WorldWideScience

Sample records for slot-scan digital chest

  1. Fundamental imaging characteristics of a slot-scan digital chest radiographic system

    International Nuclear Information System (INIS)

    Samei, Ehsan; Saunders, Robert S.; Lo, Joseph Y.; Dobbins, James T. III; Jesneck, Jonathan L.; Floyd, Carey E.; Ravin, Carl E.

    2004-01-01

    Our purpose in this study was to evaluate the fundamental image quality characteristics of a new slot-scan digital chest radiography system (ThoraScan, Delft Imaging Systems/Nucletron, Veenendaal, The Netherlands). The linearity of the system was measured over a wide exposure range at 90, 117, and 140 kVp with added Al filtration. System uniformity and reproducibility were established with an analysis of images from repeated exposures. The modulation transfer function (MTF) was evaluated using an established edge method. The noise power spectrum (NPS) and the detective quantum efficiency (DQE) of the system were evaluated at the three kilo-voltages over a range of exposures. Scatter fraction (SF) measurements were made using a posterior beam stop method and a geometrical chest phantom. The system demonstrated excellent linearity, but some structured nonuniformities. The 0.1 MTF values occurred between 3.3-3.5 mm -1 . The DQE(0.15) and DQE(2.5) were 0.21 and 0.07 at 90 kVp, 0.18 and 0.05 at 117 kVp, and 0.16 and 0.03 at 140 kVp, respectively. The system exhibited remarkably lower SFs compared to conventional full-field systems with anti-scatter grid, measuring 0.13 in the lungs and 0.43 in the mediastinum. The findings indicated that the slot-scan design provides marked scatter reduction leading to high effective DQE (DQE eff ) of the system and reduced patient dose required to achieve high image quality

  2. Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging

    International Nuclear Information System (INIS)

    Liu Xinming; Shaw, Chris C.; Lai, Chao-Jen; Wang Tianpeng

    2011-01-01

    Purpose: To investigate and compare the scatter rejection properties and low-contrast performance of the scan equalization digital radiography (SEDR) technique to the slot-scan and conventional full-field digital radiography techniques for chest imaging. Methods: A prototype SEDR system was designed and constructed with an a-Se flat-panel (FP) detector to improve image quality in heavily attenuating regions of an anthropomorphic chest phantom. Slot-scanning geometry was used to reject scattered radiation without attenuating primary x rays. The readout scheme of the FP was modified to erase accumulated scatter signals prior to image readout. A 24-segment beam width modulator was developed to regulate x-ray exposures regionally and compensate for the low x-ray flux in heavily attenuating regions. To measure the scatter-to-primary ratios (SPRs), a 2 mm thick lead plate with a 2-D array of aperture holes was used to measure the primary signals, which were then subtracted from those obtained without the lead plate to determine scatter components. A 2-D array of aluminum beads (3 mm in diameter) was used as the low-contrast objects to measure the contrast ratios (CRs) and contrast-to-noise ratios (CNRs) for evaluating the low-contrast performance in chest phantom images. A set of two images acquired with the same techniques were subtracted from each other to measure the noise levels. SPRs, CRs, and CNRs of the SEDR images were measured in four anatomical regions of chest phantom images and compared to those of slot-scan images and full-field images acquired with and without antiscatter grid. Results: The percentage reduction of SPR (percentage of SPRs reduced with scatter removal/rejection methods relative to that for nongrid full-field imaging) averaged over four anatomical regions was measured to be 80%, 83%, and 71% for SEDR, slot-scan, and full-field with grid, respectively. The average CR over four regions was found to improve over that for nongrid full

  3. Comparison of scatter rejection and low-contrast performance of scan equalization digital radiography (SEDR), slot-scan digital radiography, and full-field digital radiography systems for chest phantom imaging

    Energy Technology Data Exchange (ETDEWEB)

    Liu Xinming; Shaw, Chris C.; Lai, Chao-Jen; Wang Tianpeng [Department of Imaging Physics, Digital Imaging Research Laboratory, University of Texas M. D. Anderson Cancer Center, Houston, Texas 77030-4009 (United States)

    2011-01-15

    Purpose: To investigate and compare the scatter rejection properties and low-contrast performance of the scan equalization digital radiography (SEDR) technique to the slot-scan and conventional full-field digital radiography techniques for chest imaging. Methods: A prototype SEDR system was designed and constructed with an a-Se flat-panel (FP) detector to improve image quality in heavily attenuating regions of an anthropomorphic chest phantom. Slot-scanning geometry was used to reject scattered radiation without attenuating primary x rays. The readout scheme of the FP was modified to erase accumulated scatter signals prior to image readout. A 24-segment beam width modulator was developed to regulate x-ray exposures regionally and compensate for the low x-ray flux in heavily attenuating regions. To measure the scatter-to-primary ratios (SPRs), a 2 mm thick lead plate with a 2-D array of aperture holes was used to measure the primary signals, which were then subtracted from those obtained without the lead plate to determine scatter components. A 2-D array of aluminum beads (3 mm in diameter) was used as the low-contrast objects to measure the contrast ratios (CRs) and contrast-to-noise ratios (CNRs) for evaluating the low-contrast performance in chest phantom images. A set of two images acquired with the same techniques were subtracted from each other to measure the noise levels. SPRs, CRs, and CNRs of the SEDR images were measured in four anatomical regions of chest phantom images and compared to those of slot-scan images and full-field images acquired with and without antiscatter grid. Results: The percentage reduction of SPR (percentage of SPRs reduced with scatter removal/rejection methods relative to that for nongrid full-field imaging) averaged over four anatomical regions was measured to be 80%, 83%, and 71% for SEDR, slot-scan, and full-field with grid, respectively. The average CR over four regions was found to improve over that for nongrid full

  4. Benefits and unexpected artifacts of biplanar digital slot-scanning imaging in children

    Energy Technology Data Exchange (ETDEWEB)

    Blumer, Steven L. [Nemours/A.I duPont Hospital for Children, Department of Medical Imaging, Wilmington, DE (United States); Dinan, David [Nemours Children' s Hospital, Orlando, FL (United States); Grissom, Leslie E. [Nemours/Alfred I. duPont Hospital for Children, Department of Radiology, Wilmington, DE (United States)

    2014-07-15

    Biplanar digital slot-scanning allows for relatively low-dose orthopedic imaging, an advantage in imaging children given the growing concerns regarding radiosensitivity. We have used this system for approximately 1 year for orthopedic imaging of the spine and lower extremities. We have noted advantages of using the digital slot-scanning system when compared with computed radiographic and standard digital radiographic imaging systems, but we also found unexpected but common imaging artifacts that are the direct result of the imaging method and that have not been reported. This pictorial essay serves to familiarize radiologists with the advantages of the digital slot-scanning system as well as imaging artifacts common with this new technology. (orig.)

  5. A CdZnTe slot-scanned detector for digital mammography

    International Nuclear Information System (INIS)

    Mainprize, James G.; Ford, Nancy L.; Yin Shi; Gordon, Eli E.; Hamilton, William J.; Tuemer, Tuemay O.; Yaffe, Martin J.

    2002-01-01

    A new high-resolution detector has been developed for use in a slot-scanned digital mammography system. The detector is a hybrid device that consists of a CCD operating in time-delay integration mode that is bonded to a 150-μm-thick CdZnTe photoconductor array. The CCD was designed with a detector element pitch of 50 μm. Two devices were evaluated with differing crystalline quality. Incomplete charge collection was a source of reduction in DQE. This occurs in both devices due to characteristically low mobility-lifetime products for CdZnTe, with the greatest losses demonstrated by the multicrystalline sample. The mobility-lifetime products for the multicrystalline device were found to be 2.4x10 -4 and 4.0x10 -7 cm 2 /V for electrons and holes, respectively. The device constructed with higher quality single crystal CdZnTe demonstrated mobility-lifetime products of 1.0x10 -4 and 4.4x10 -6 cm 2 /V for electrons and holes. The MTF and DQE for the device were measured at several exposures and results were compared to predictions from a linear systems model of signal and noise propagation. The MTF at a spatial frequency of 10 mm -1 exceeded 0.18 and 0.56 along the scan and slot directions, respectively. Scanning motion and CCD design limited the resolution along the scan direction. For an x-ray beam from a tungsten target tube with 40 μm molybdenum filtration operated at 26 kV, the single crystal device demonstrated a DQE(0) of 0.70±0.02 at 7.1x10 -6 C/kg (27 mR) exposure to the detector, despite its relatively poor charge collection efficiency

  6. Digital radiography of the chest

    International Nuclear Information System (INIS)

    Sakurai, Kenji; Hachiya, Junichi; Korenaga, Tateo; Nitatori, Toshiaki; Miyasaka, Yasuo; Furuya, Yoshiro

    1984-01-01

    Initial clinical experience in digital chest radiography utilizing photostimulable phosphor and scanning laser stimulated luminescence was reported. Image quality of conventional film/screen radiography and digital radiography was compared in 30 normal cases. Reflecting wide dynamic range of the system, improved image quality was confirmed in all 30 cases, particularly in visibility of various mediastinal structures and pulmonary vessels. High sensor sensitivity of the system enabled digital radiography to reduce radiation dose requirement significantly. Diagnostically acceptable chest images were obtained with approximately 1/5 of routine dose for conventional radiography without significant image quality degradation. Some artifact created by digital processing were mostly overcome by a routine use of simultaneous display of two different types of image processing and therefore was not an actual drawback from diagnostic standpoint. Further technical advancement of the system to be seen for digital storage, retrieval and tranceference of images. (author)

  7. Digital chest radiography

    DEFF Research Database (Denmark)

    Debess, Jeanne Elisabeth; Johnsen, Karen Kirstine; Thomsen, Henrik

    on collimation and dose reduction in digital chest radiography Methods and Materials A retrospective study of digital chest radiography is performed to evaluate the primary x-ray tube collimation of the PA and lateral radiographs. Data from one hundred fifty self-reliant female patients between 15 and 55 years...

  8. Early appearance of SARS on chest CT scan

    International Nuclear Information System (INIS)

    Cheng Xiaoguang; Feng Suchen; Xia Guoguang; Zhao Tao; Gu Xiang; Qu Hui

    2003-01-01

    Objective: To evaluate the early appearance of SARS on chest CT scan and its role in the early diagnosis. Methods: Forty cases of SARS in keeping with the criteria of the Ministry of Health had chest CT scans within 7 days of onset of symptoms, and CR chest X-ray films were available as well. These chest X-rays and CT images were retrospectively reviewed to determine if there were any abnormalities on the images. The lesions on the chest CT images were then further analyzed in terms of the number, location, size, and density. Results: Positive abnormalities on chest CT scans were revealed in all 40 SARS cases. Positive findings on CR chest films were showed in only 25 cases, equivocal in 6, and normal in 9 cases. The main abnormalities seen on CT and X-rays were pulmonary infiltrations varied markedly in severity. 70 % cases had 1 or 2 lesions on chest CT scan, 30 % cases had 3 or more lesions. The lesions seen on chest CT scan tended to be ground-glass opacification, sometimes with consolidation which was very faint and inhomogeneous, easily missed on chest X-rays. Typically the lesions were located in the periphery of the lung, or both central and peripheral lung, but very rare in a pure central location. They were commonly in the shape of patch or ball. Conclusions: Chest CT scan is much more sensitive in detecting the lesions of the lung in SARS. The early appearance of SARS on chest CT scan is characteristic but non-specific, indicating that chest CT scan plays a very important role in the early diagnosis and differential diagnosis of SARS

  9. Scan equalization digital radiography (SEDR) implemented with an amorphous selenium flat-panel detector: initial experience

    Energy Technology Data Exchange (ETDEWEB)

    Liu Xinming; Lai Chaojen; Chen Lingyun; Han Tao; Zhong Yuncheng; Shen Youtao; Wang Tianpeng; Shaw, Chris C [Department of Imaging Physics, Digital Imaging Research Laboratory, University of Texas M. D. Anderson Cancer Center, Houston, TX 77030-4009 (United States)], E-mail: xliu@di.mdacc.tmc.edu

    2009-11-21

    It is well recognized in projection radiography that low-contrast detectability suffers in heavily attenuating regions due to excessively low x-ray fluence to the image receptor and higher noise levels. Exposure equalization can improve image quality by increasing the x-ray exposure to heavily attenuating regions, resulting in a more uniform distribution of exposure to the detector. Image quality is also expected to be improved by using the slot-scan geometry to reject scattered radiation effectively without degrading primary x-rays. This paper describes the design of a prototype scan equalization digital radiography (SEDR) system implemented with an amorphous silicon (a-Si) thin-film transistor (TFT) array-based flat-panel detector. With this system, slot-scan geometry with alternate line erasure and readout (ALER) technique was used to achieve scatter rejection. A seven-segment beam height modulator assembly was mounted onto the fore collimator to regulate exposure regionally for chest radiography. The beam modulator assembly, consisting of micro linear motors, lead screw cartridge with lead (Pb) beam blockers attached, position feedback sensors and motor driver circuitry, has been tested and found to have an acceptable response for exposure equalization in chest radiography. An anthropomorphic chest phantom was imaged in the posterior-anterior (PA) view under clinical conditions. Scatter component, primary x-rays, scatter-to-primary ratios (SPRs) and primary signal-to-noise ratios (PSNRs) were measured in the SEDR images to evaluate the rejection and redistribution of scattered radiation, and compared with those for conventional full-field imaging with and without anti-scatter grid methods. SPR reduction ratios (SPRRRs, defined as the differences between the non-grid full-field SPRs and the reduced SPRs divided by the former) yielded approximately 59% for the full-field imaging with grid and 82% for the SEDR technique in the lungs, and 77% for the full

  10. Lung cancer detection with digital chest tomosynthesis: first round results from the SOS observational study.

    Science.gov (United States)

    Bertolaccini, Luca; Viti, Andrea; Tavella, Chiara; Priotto, Roberto; Ghirardo, Donatella; Grosso, Maurizio; Terzi, Alberto

    2015-04-01

    Baseline results of the Studio OSservazionale (SOS), observational study, a single-arm observational study of digital chest tomosynthesis for lung cancer detection in an at-risk population demonstrated a detection rate of lung cancer comparable to that of studies that used low dose CT scan (LDCT). We present the results of the first round. Totally 1,703 out of 1,843 (92%) subjects who had a baseline digital chest tomosynthesis underwent a first round reevaluation after 1 year. At first round chest digital tomosynthesis, 13 (0.7%) subjects had an indeterminate nodule larger than 5 mm and underwent low-dose CT scan for nodule confirmation. PET/CT study was obtained in 10 (0.5%) subjects and 2 subjects had a low-dose CT follow up. Surgery, either video-assisted thoracoscopic or open surgery for indeterminate pulmonary nodules was performed in 10 (0.2%) subjects. A lung cancer was diagnosed and resected in five patients. The lung cancer detection rate at first round was 0.3% (5/1,703). The detection rate of lung cancer at first round for tomosynthesis is comparable to rates reported for CT. In addition, results of first round digital chest tomosynthesis confirm chest tomosynthesis as a possible first-line lung cancer-screening tool.

  11. Digital tomosynthesis of the chest: A literature review

    International Nuclear Information System (INIS)

    Molk, N.; Seeram, E.

    2015-01-01

    Digital tomosynthesis is a relatively novel imaging modality using limited angle tomography to provide 3D imaging. The purpose of this review is to compare the sensitivity of digital tomosynthesis of the chest and plain film chest imaging in accurately identifying pulmonary nodules and to compare the effective dose between standard chest examinations using digital tomosynthesis and CT. A review of current literature has shown that small scale studies found digital tomosynthesis to be three times more effective in identifying pulmonary nodules compared to conventional radiography and at lower doses compared with routine chest CT examinations. This indicates that tomosynthesis could potentially be a beneficial imaging modality and could be used in a number of ways to detect and monitor pulmonary nodules for cancer. However with limited research, large-scale studies would need to be performed to confirm its benefits and identify where it is best used in the clinical setting. - Highlights: • The detection of pulmonary nodules is compared between tomosynthesis and plain film. • The effective dose of digital chest tomosynthesis and chest CT are compared. • The place of digital tomosynthesis of the chest in the clinical setting is explored. • Three times more pulmonary nodules are seen with tomosynthesis. • The effective dose of tomosynthesis is significantly lower than CT

  12. Image processing in digital chest radiography

    International Nuclear Information System (INIS)

    Manninen, H.; Partanen, K.; Lehtovirta, J.; Matsi, P.; Soimakallio, S.

    1992-01-01

    The usefulness of digital image processing of chest radiographs was evaluated in a clinical study. In 54 patients, chest radiographs in the posteroanterior projection were obtained by both 14 inch digital image intensifier equipment and the conventional screen-film technique. The digital radiographs (512x512 image format) viewed on a 625 line monitor were processed in 3 different ways: 1.standard display; 2.digital edge enhancement for the standard display; 3.inverse intensity display. The radiographs were interpreted independently by 3 radiologists. Diagnoses were confirmed by CT, follow-up radiographs and clinical records. Chest abnormalities of the films analyzed included 21 primary lung tumors, 44 pulmonary nodules, 16 cases with mediastinal disease, 17 with pneumonia /atelectasis. Interstitial lung disease, pleural plaques, and pulmonary emphysema were found in 30, 18 and 19 cases respectively. Sensitivity of conventional radiography when averaged overall findings was better than that of digital techniques (P<0.001). Differences in diagnostic accuracy measured by sensitivity and specificity between the 3 digital display modes were small. Standard image display showed better sensitivity for pulmonary nodules (0.74 vs 0.66; P<0.05) but poorer specificity for pulmonary emphysema (0.85 vs 0.93; P<0.05) compared with inverse intensity display. It is concluded that when using 512x512 image format, the routine use of digital edge enhancement and tone reversal at digital chest radiographs is not warranted. (author). 12 refs.; 4 figs.; 2 tabs

  13. Digital radiography of the chest: state of the art

    International Nuclear Information System (INIS)

    Souto, M.; Malagari, K.S.; Tucker, D.; Tahoces, P.G.; Correa, J.; Benakis, V.S.; Roussos, C.; Strigaris, K.A.; Vidal, J.J.; Barnes, G.T.; Fraser, R.G.

    1994-01-01

    Digital image acquisition possesses a number of advantages over conventional systems in radiographic examination of the chest, the most important of which is its greater dynamic range. In addition, once digital images are acquired, they can be processed by computer in ways that cannot be rivalled by conventional analog techniques. Finally, digital images can be stored, retrieved and transmitted to local or remote sites. Here the status of the different digital systems employed in chest radiology and commonly used image processing techniques are reviewed. Also discussed are the current clinical applications of integrating digital chest radiography with a picture archiving and communication system (PACS) along with the difficulties typically encountered. Studies with a variety of digital techniques have been carried out on several fronts. Computed radiography based on photostimulable phosphor (CR) has replaced screen-film imaging in certain applications (i.e. bedside imaging). However, CR has limitations, namely its poor X-ray utilisation efficiency at high X-ray tube voltages and sensitivity to scatter; therefore, it is not ideal for all applications. Recently, a dedicated digital chest unit with excellent X-ray utilisation efficiency at high X-ray tube potentials has been introduced. On the basis of the state-of-the-art capabilities and research during the past decade, recommendations are made regarding the most desirable equipment specifications for dedicated and bedside digital chest radiography. (orig.)

  14. Image quality of a Konica Regius 336 digital system in chest radiography

    International Nuclear Information System (INIS)

    Ostinelli, A.; Frigerio, M.; Monti, A.F.; Gelosa, S.; Tognoli, P.; Perniola, N.; Gozzi, G.

    2000-01-01

    Digital radiographic systems permit to optimize execution, depiction and storage of radiological images. Since a Regius 336 digital system (Konica Corp. Tokyo, Japan) devoted to chest radiography Department of S. Anna Hospital in Como, Italy, it was investigated its performance relative to image quality. Konica Regius 336 is a computed radiography system made of a phosphorescence detector plate which is scanned with an infrared semiconductor laser beam. The radiographic image obtained from the detector is subjected to image processing, which allows a stable output and the nonlinear curve typical of conventional radiographic systems. Image quality was assessed based on the following parameters: dose, contrast, noise and spatial resolution. As reference, it was assessed the same parameters on a Cronex 88 analogic chest-changer (DuPont Pharma, North Billerica, Mass, USA). The Regius 336 air kerma values were always higher than the analogic ones (about 10%), both with and without a chest phantom; noise was also greater than in analogic images, sometimes even doubled. The optical densities of a step wedge and the spatial resolution of the digital chest-changer are independent of the X-ray tube voltage consequent to broader optical latitude. Inversely, the analogic images of the wedges show great optical density variability as a function of the X-ray tube voltage (in a range of 2). The modulation transfer functions of the two systems have the same trend. The performance of the Konica Regius 336 is nearly equivalent to that of an analogic system. The main advantages of the digital system are a standard output, lower consumption of radiographic films, higher productiveness and better image quality standard level [it

  15. [Development of a digital chest phantom for studies on energy subtraction techniques].

    Science.gov (United States)

    Hayashi, Norio; Taniguchi, Anna; Noto, Kimiya; Shimosegawa, Masayuki; Ogura, Toshihiro; Doi, Kunio

    2014-03-01

    Digital chest phantoms continue to play a significant role in optimizing imaging parameters for chest X-ray examinations. The purpose of this study was to develop a digital chest phantom for studies on energy subtraction techniques under ideal conditions without image noise. Computed tomography (CT) images from the LIDC (Lung Image Database Consortium) were employed to develop a digital chest phantom. The method consisted of the following four steps: 1) segmentation of the lung and bone regions on CT images; 2) creation of simulated nodules; 3) transformation to attenuation coefficient maps from the segmented images; and 4) projection from attenuation coefficient maps. To evaluate the usefulness of digital chest phantoms, we determined the contrast of the simulated nodules in projection images of the digital chest phantom using high and low X-ray energies, soft tissue images obtained by energy subtraction, and "gold standard" images of the soft tissues. Using our method, the lung and bone regions were segmented on the original CT images. The contrast of simulated nodules in soft tissue images obtained by energy subtraction closely matched that obtained using the gold standard images. We thus conclude that it is possible to carry out simulation studies based on energy subtraction techniques using the created digital chest phantoms. Our method is potentially useful for performing simulation studies for optimizing the imaging parameters in chest X-ray examinations.

  16. A Monte Carlo-based model for simulation of digital chest tomo-synthesis

    International Nuclear Information System (INIS)

    Ullman, G.; Dance, D. R.; Sandborg, M.; Carlsson, G. A.; Svalkvist, A.; Baath, M.

    2010-01-01

    The aim of this work was to calculate synthetic digital chest tomo-synthesis projections using a computer simulation model based on the Monte Carlo method. An anthropomorphic chest phantom was scanned in a computed tomography scanner, segmented and included in the computer model to allow for simulation of realistic high-resolution X-ray images. The input parameters to the model were adapted to correspond to the VolumeRAD chest tomo-synthesis system from GE Healthcare. Sixty tomo-synthesis projections were calculated with projection angles ranging from + 15 to -15 deg. The images from primary photons were calculated using an analytical model of the anti-scatter grid and a pre-calculated detector response function. The contributions from scattered photons were calculated using an in-house Monte Carlo-based model employing a number of variance reduction techniques such as the collision density estimator. Tomographic section images were reconstructed by transferring the simulated projections into the VolumeRAD system. The reconstruction was performed for three types of images using: (i) noise-free primary projections, (ii) primary projections including contributions from scattered photons and (iii) projections as in (ii) with added correlated noise. The simulated section images were compared with corresponding section images from projections taken with the real, anthropomorphic phantom from which the digital voxel phantom was originally created. The present article describes a work in progress aiming towards developing a model intended for optimisation of chest tomo-synthesis, allowing for simulation of both existing and future chest tomo-synthesis systems. (authors)

  17. Analysis the findings of chest radiograph and CT scan in 217 acute thoracic trauma

    International Nuclear Information System (INIS)

    Huang Shaoying

    2005-01-01

    Objective: To evaluate chest radiograph and CT scan in assessing acute thoracic trauma. Methods: Retrospectively analyzed the findings of chest radiograph and CT scan in 217 cases of acute thoracic trauma and positive rate of each modality was compared. Results: The positive rate of rib and clavicle fracture was higher in chest radiograph than these in CT scan. But the positive rate of chest wall hematoma, mediastinum and subcutaneous emphysema, pneumothorax, hydropneumothorax, damage of lung parenchyma and traumatic pulmonary atelectasis was higher in CT scan than those in chest radiograph. Conclusion: The application of the combined imaging modalities improves assessing of acute thoracic trauma, because the imaging manifestation of the lesion is various. (authors)

  18. Diagnostic image quality of video-digitized chest images

    International Nuclear Information System (INIS)

    Winter, L.H.; Butler, R.B.; Becking, W.B.; Warnars, G.A.O.; Haar Romeny, B. ter; Ottes, F.P.; Valk, J.-P.J. de

    1989-01-01

    The diagnostic accuracy obtained with the Philips picture archiving and communications subsystem was investigated by means of an observer performance study using receiver operating characteristic (ROC) analysis. The image qualities of conventional films and video digitized images were compared. The scanner had a 1024 x 1024 x 8 bit memory. The digitized images were displayed on a 60 Hz interlaced display monitor 1024 lines. Posteroanterior (AP) roetgenograms of a chest phantom with superimposed simulated interstitial pattern disease (IPD) were produced; there were 28 normal and 40 abnormal films. Normal films were produced by the chest phantom alone. Abnormal films were taken of the chest phantom with varying degrees of superimposed simulated intersitial disease (PND) for an observer performance study, because the results of a simulated interstitial pattern disease study are less likely to be influenced by perceptual capabilities. The conventional films and the video digitized images were viewed by five experienced observers during four separate sessions. Conventional films were presented on a viewing box, the digital images were displayed on the monitor described above. The presence of simulated intersitial disease was indicated on a 5-point ROC certainty scale by each observer. We analyzed the differences between ROC curves derived from correlated data statistically. The mean time required to evaluate 68 digitized images is approximately four times the mean time needed to read the convential films. The diagnostic quality of the video digitized images was significantly lower (at the 5% level) than that of the conventional films (median area under the curve (AUC) of 0.71 and 0.94, respectively). (author). 25 refs.; 2 figs.; 4 tabs

  19. Ultrasound detection of pneumothorax compared with chest X-ray and computed tomography scan.

    Science.gov (United States)

    Nagarsheth, Khanjan; Kurek, Stanley

    2011-04-01

    Pneumothorax after trauma can be a life threatening injury and its care requires expeditious and accurate diagnosis and possible intervention. We performed a prospective, single blinded study with convenience sampling at a Level I trauma center comparing thoracic ultrasound with chest X-ray and CT scan in the detection of traumatic pneumothorax. Trauma patients that received a thoracic ultrasound, chest X-ray, and chest CT scan were included in the study. The chest X-rays were read by a radiologist who was blinded to the thoracic ultrasound results. Then both were compared with CT scan results. One hundred and twenty-five patients had a thoracic ultrasound performed in the 24-month period. Forty-six patients were excluded from the study due to lack of either a chest X-ray or chest CT scan. Of the remaining 79 patients there were 22 positive pneumothorax found by CT and of those 18 (82%) were found on ultrasound and 7 (32%) were found on chest X-ray. The sensitivity of thoracic ultrasound was found to be 81.8 per cent and the specificity was found to be 100 per cent. The sensitivity of chest X-ray was found to be 31.8 per cent and again the specificity was found to be 100 per cent. The negative predictive value of thoracic ultrasound for pneumothorax was 0.934 and the negative predictive value for chest X-ray for pneumothorax was found to be 0.792. We advocate the use of chest ultrasound for detection of pneumothorax in trauma patients.

  20. Automated image quality assessment for chest CT scans.

    Science.gov (United States)

    Reeves, Anthony P; Xie, Yiting; Liu, Shuang

    2018-02-01

    Medical image quality needs to be maintained at standards sufficient for effective clinical reading. Automated computer analytic methods may be applied to medical images for quality assessment. For chest CT scans in a lung cancer screening context, an automated quality assessment method is presented that characterizes image noise and image intensity calibration. This is achieved by image measurements in three automatically segmented homogeneous regions of the scan: external air, trachea lumen air, and descending aorta blood. Profiles of CT scanner behavior are also computed. The method has been evaluated on both phantom and real low-dose chest CT scans and results show that repeatable noise and calibration measures may be realized by automated computer algorithms. Noise and calibration profiles show relevant differences between different scanners and protocols. Automated image quality assessment may be useful for quality control for lung cancer screening and may enable performance improvements to automated computer analysis methods. © 2017 American Association of Physicists in Medicine.

  1. The Beatles, the Nobel Prize, and CT scanning of the chest.

    Science.gov (United States)

    Goodman, Lawrence R

    2010-01-01

    From its first test scan on a mouse, in 1967, to current medical practice, the CT scanner has become a core imaging tool in thoracic diagnosis. Initially financed by money from Beatles' record sales, the first patient scan was performed in 1971. Only 8 years later, a Nobel Prize in Physics and Medicine was awarded to Hounsfield and Cormack for their discovery. This article traces the history of CT scanner development and how each technical advance expanded chest diagnostic frontiers. Chest imaging now accounts for 30% of all CT scanning.

  2. Utility of CT scan for the diagnosis of chest wall tuberculosis

    Energy Technology Data Exchange (ETDEWEB)

    Khalil, A.; Le Breton, C.; Tassart, M.; Korzec, J.; Bigot, J.M.; Carette, M.F. [Department of Radiology, Tenon Hospital, Paris (France)

    1999-10-01

    The objective of this study was to determine the utility of CT scan findings for the diagnosis of chest wall tuberculosis, excluding the spine. We reviewed 15 patients (13 Africans and 2 Indians) with chest wall tuberculosis, retrospectively. The radiologic examination consisted of a plain X-ray and a CT scan of the chest for each patient. The site of disease was the rib in 13 patients or the body of the sternum in 2 patients. One rib was involved in 11 patients, 2 contiguous ribs (one site) in 2 patients, and bilateral disease (two sites) was observed in the remaining patient. The 14 rib sites involved the posterior arc or costovertebral joint in 11 cases, the anterior arc in 2 cases, and the anterior and middle arc in 1 case. The CT scan findings were an abscess (n = 14) or a soft tissue mass (n = 2), osteolytic lesions (n = 13), periosteal reaction (n = 10), and sequestrum (n = 14). Bone sclerosis was observed only in 3 cases of rib involvement. The association of a soft tissue abscess, an osteolytic lesion, and sequestrum, especially in immigrants to France, suggests chest wall tuberculosis on CT scan. (orig.) With 5 figs., 2 tabs., 11 refs.

  3. Digital luminescence radiography of the chest

    International Nuclear Information System (INIS)

    Kehler, M.

    1991-10-01

    The aim of the present study was to evaluate the efficacy of a digital system in chest radiology compared to the conventional film-screen system. The first studies (1-3) were purely clinical, had two parts, one clinical and one using phantoms, and the 5:th used solely phantoms. Except for the first - pilot - study, the studies were performed as receiver operating characteristic (ROC) analysis. From one exposure, two digital radiographs were obtained, one simulating the film-screen radiograph and one enhanced, using an unsharp mask. The conventional radiograph was compared to this double-image, but in addition to this, even to the simulated normal and enhanced separately (1-3). To evaluate the value of inverted (positive) radiographs, the original digital (negative) radiographs were inverted, and then compared to the originals (4). As digitzation means easy storing and transfer of data and possibility of electronic display, the diagnostic performance of an interactive workstation was assessed (5). In the clinical studies, a variety of chest affections were used: atelectasis, tumor, pneumothorax, fibrosis, mediastinal and bony changes, tuberculosis, incompensations and enlargement of the heart (1), pneumothorax (2), fibrosis (3), and tumor (4). In the phantom studies, test objects simulating tumors (4) and pneumothorax (5) were used. In no study was statistical significant difference seen between the digital and conventional system (p>0.05). Neither in the clinical nor the phantom study did inversion of the radiographs improve diagnostic performance. The workstation performed almost equally well as the radiographs even with a resolution of 1.25 1p/mm compared to the digital radiographs 2.5 and film-screen radiographs 5 1p/mm. (au) (50 refs.)

  4. Digitizing pediatric chest radiographs: comparison of low-cost, commercial off-the-shelf technologies

    International Nuclear Information System (INIS)

    Ruess, L.; Shiels, K.C.; Cho, K.H.; O'Connor, S.C.; Uyehara, C.F.T.; Person, D.A.; Whitton, R.K.

    2001-01-01

    Objective: To compare low-cost, off-the-shelf technology for digitizing pediatric chest radiographs. Materials and methods: Forty pediatric chest radiographs (hard copy), each with a single abnormality, were digitized using a commercial film digitizer and two low-cost methods: a digital camera and a flatbed scanner. A stratified, randomized, block design was used where 20 readers evaluated 40 different images to determine the ability to accurately detect the abnormality. Readers then rated all 160 images (40 images x 4 methods) for conspicuity of the abnormality and overall image quality. Results: Abnormalities were correctly identified on 82.3 % of hard copy images, 82.9 % of flatbed scanner images, 74.3 % of film digitizer images, and 69.7 % of digital camera images (p flatbed scanner > film digitizer > digital camera images. Conclusion: A low-cost flatbed scanner yielded digital pediatric chest images which were significantly superior to digital camera images While flatbed scanner images were interpreted with the equivalent diagnostic accuracy of hard copy images, they were rated lower for image quality and lesion conspicuity. (orig.)

  5. Digital chest radiography system with amorphous selenium flat-panel detectors: qualitative and dosimetric comparison with a dedicated film-screen system

    International Nuclear Information System (INIS)

    Prato, Antonio; Fava Cesare; Ropolo, Roberto

    2005-01-01

    Purpose. To compare the quality and radiation dose of a conventional film-screen system and a digital system with amorphous selenium detectors in the study of the chest, by verifying overall performance and exposure levels for the main chest structures in patients of different sizes. Materials and methods. An analogic system (Chest Changer, Dupont, Day-light model 1000) and a digital system (Directray Rad 1000C, Hologic) were tested on a total of 1000 patients randomly assigned to one of two groups of 500 subjects.The patients were further subdivided according to BMI (Body Mass Index). Image quality was determined by two chest radiologists who evaluated eight anatomical structures. The entrance surface dose (skin-dose), calculated based on the exposure parameters, was taken as the patient dose. Results. Mean dose delivered was very similar for both techniques in the PA view (0.28 mGy), but it was greater in the LL projections obtained with the digital system (1.20 rnGy versus 0.83 mGy). The highest overall scores were assigned to 43% and 23.2% analogic radiograms and 64% and 70.2% digital radiograms, for the PA and LL projections respectively. The scores assigned to the various anatomical structures confirmed the better performance of the digital system in almost all of the regions considered. Conclusions. The mean quality of radiograms is definitely higher with the digital system, in particular in the LL projections, where the higher patient doses are counterbalanced by fewer repeated scans. The greater level of exposure in the digital system appears nonetheless tolerable on account of the greater informativeness and therefore diagnostic gain and also considering the possibilities for improving the system [it

  6. Nodule detection in digital chest radiography: Summary of the radius chest trial

    International Nuclear Information System (INIS)

    Haakansson, M.; Baath, M.; Boerjesson, S.; Kheddache, S.; Grahn, A.; Ruschin, M.; Tingberg, A.; Mattson, S.; Maansson, L. G.

    2005-01-01

    As a part of the Europe-wide research project 'Unification of physical and clinical requirements for medical X-ray imaging' - governed by the Radiological Imaging Unification Strategies (RADIUS) Group - a major image quality trial was conducted by members of the group. The RADIUS chest trial aimed at thoroughly examining various aspects of nodule detection in digital chest radiography, such as the effects of nodule location, system noise, anatomical noise, and anatomical background. The main findings of the RADIUS chest trial concerning the detection of a lung nodule with a size in the order of 10 mm can be summarised as: (1) the detectability of the nodule is largely dependent on its location in the chest, (2) the system noise has a minor impact on the detectability at the dose levels used today, (3) the disturbance of the anatomical noise is larger than that of the system noise but smaller than that of the anatomical background and (4) the anatomical background acts as noise to a large extent and is the major image component affecting the detectability of the nodule. (authors)

  7. Anteroposterior chest radiograph vs. chest CT scan in early detection of pneumothorax in trauma patients

    Directory of Open Access Journals (Sweden)

    Omar Hesham R

    2011-09-01

    Full Text Available Abstract Pneumothorax is a common complication following blunt chest wall trauma. In these patients, because of the restrictions regarding immobilization of the cervical spine, Anteroposterior (AP chest radiograph is usually the most feasible initial study which is not as sensitive as the erect chest X-ray or CT chest for detection of a pneumothorax. We will present 3 case reports which serve for better understanding of the entity of occult pneumothorax. The first case is an example of a true occult pneumothorax where an initial AP chest X-ray revealed no evidence of pneumothorax and a CT chest immediately performed revealed evidence of pneumothorax. The second case represents an example of a missed rather than a truly occult pneumothorax where the initial chest radiograph revealed clues suggesting the presence of pneumothorax which were missed by the reading radiologist. The third case emphasizes the fact that "occult pneumothorax is predictable". The presence of subcutaneous emphesema and pulmonary contusion should call for further imaging with CT chest to rule out pneumothorax. Thoracic CT scan is therefore the "gold standard" for early detection of a pneumothorax in trauma patients. This report aims to sensitize readers to the entity of occult pneumothorax and create awareness among intensivists and ER physicians regarding the proper diagnosis and management.

  8. Anteroposterior chest radiograph vs. chest CT scan in early detection of pneumothorax in trauma patients.

    Science.gov (United States)

    Omar, Hesham R; Mangar, Devanand; Khetarpal, Suneel; Shapiro, David H; Kolla, Jaya; Rashad, Rania; Helal, Engy; Camporesi, Enrico M

    2011-09-27

    Pneumothorax is a common complication following blunt chest wall trauma. In these patients, because of the restrictions regarding immobilization of the cervical spine, Anteroposterior (AP) chest radiograph is usually the most feasible initial study which is not as sensitive as the erect chest X-ray or CT chest for detection of a pneumothorax. We will present 3 case reports which serve for better understanding of the entity of occult pneumothorax. The first case is an example of a true occult pneumothorax where an initial AP chest X-ray revealed no evidence of pneumothorax and a CT chest immediately performed revealed evidence of pneumothorax. The second case represents an example of a missed rather than a truly occult pneumothorax where the initial chest radiograph revealed clues suggesting the presence of pneumothorax which were missed by the reading radiologist. The third case emphasizes the fact that "occult pneumothorax is predictable". The presence of subcutaneous emphesema and pulmonary contusion should call for further imaging with CT chest to rule out pneumothorax. Thoracic CT scan is therefore the "gold standard" for early detection of a pneumothorax in trauma patients. This report aims to sensitize readers to the entity of occult pneumothorax and create awareness among intensivists and ER physicians regarding the proper diagnosis and management.

  9. Practical applications of digital tomosynthesis of the chest

    International Nuclear Information System (INIS)

    Galea, A.; Durran, A.; Adlan, T.; Gay, D.; Riordan, R.; Dubbins, P.; Williams, M.P.

    2014-01-01

    Digital tomosynthesis is a radiographic technique that generates a number of coronal raw images of a patient from a single pass of the x-ray tube. Tomosynthesis provides some of the tomographic benefits of computed tomography (CT), but at a much lower dose of radiation and cost when compared to CT. This review illustrates the range of practical applications of digital tomosynthesis of the chest

  10. Digital training platform for interpreting radiographic images of the chest.

    Science.gov (United States)

    McLaughlin, L; Woznitza, N; Cairns, A; McFadden, S L; Bond, R; Hughes, C M; Elsayed, A; Finlay, D; McConnell, J

    2018-05-01

    Time delays and errors exist which lead to delays in patient care and misdiagnosis. Reporting clinicians follow guidance to form their own search strategy. However, little research has tested these training guides. With the use of eye tracking technology and expert input we developed a digital training platform to be used in chest image interpretation learning. Two sections of a digital training platform were planned and developed; A) a search strategy training tool to assist reporters during their interpretation of images, and B) an educational tool to communicate the search strategies of expert viewers to trainees by using eye tracking technology. A digital training platform for use in chest image interpretation was created based on evidence within the literature, expert input and two search strategies previously used in clinical practice. Images and diagrams, aiding translation of the platform content, were incorporated where possible. The platform is structured to allow the chest image interpretation process to be clear, concise and methodical. A search strategy was incorporated within the tool to investigate its use, with the possibility that it could be recommended as an evidence based approach for use by reporting clinicians. Eye tracking, a checklist and voice recordings have been combined to form a multi-dimensional learning tool, which has never been used in chest image interpretation learning before. The training platform for use in chest image interpretation learning has been designed, created and digitised. Future work will establish the efficacy of the developed approaches. Crown Copyright © 2018. Published by Elsevier Ltd. All rights reserved.

  11. Comparison of the quality of the chest film between digital radiography and conventional high kV radiography

    International Nuclear Information System (INIS)

    Zeng Qingsi; Cen Renli; Chen Ling; He Jianxun; Lin Hanfei

    2003-01-01

    Objective: To evaluate the quality and usefulness of direct digital radiography system in roentgenogram of chest in clinical practice. Methods: 1000 cases of chest roentgenograms with digital radiography and high kV conventional radiography were selected for analysis by 3 senior radiologists. Results: 1. With digital radiography system, the quality of chest film was assessed as grade A in 50.6%, grade B in 38.5%, grade C in 10.9%, and no waste film. 2. With conventional high kV radiography, the quality of chest film was assessed as grade A in 41.1%, grade B in 44.1%, grade C in 13.3%, and waste film in 1.5%. The direct digital radiography was statistically superior to the conventional high kV radiography. 3. The fine structure of the lungs could be revealed in 100.0% of chest roentgenogram with direct digital radiograph system, which was significantly higher than that acquired with the conventional high KV radiography (78.6%, P < 0.001). Conclusion: Direct digital radiography could provide the chest film with better quality than that with the conventional high kV radiography. The direct digital radiography system is easy to operate, fast in capturing imaging and could provide post-processing techniques, which will facilitate the accurate diagnosis of chest radiography

  12. Thallium myocardial scanning in the emergency department evaluation of chest pain

    International Nuclear Information System (INIS)

    Mace, S.E.

    1989-01-01

    Chest pain is a common complaint of patients seen in the emergency department. The causes are legion, and range from the non-life threatening to the potentially catastrophic. Thallium heart scanning was done prospectively in 20 patients with a ''classic'' history for myocardial infarction (eight patients) or atypical chest pain and/or associated symptoms plus an abnormal ECG (12 patients) to discern a subset of patients from whom thallium scintography may be indicated in the emergency department. Although further investigation is needed, our preliminary study suggests that myocardial scanning with thallium can be a safe, fairly rapid, and useful objective parameter in the emergency department detection of suspected myocardial infarction, and in differential diagnosis of chest pain when other data such as the history, physical examination, ECG, or enzymes are inconclusive

  13. Direct digital acquisition of neonatal portable chest radiographs

    International Nuclear Information System (INIS)

    Cohen, M.D.; Cory, D.A.; Broderick, N.J.; Smith, J.A.

    1987-01-01

    The Philips computerized radiography system utilizes a phosphor plate instead of an x-ray film to acquire radiographic images. The latent image on the plate is converted to a digital format. The authors report their initial experience with the system in more than 300 studies on intensive care neonates. The digital images provide very uniform image density. Tubes and catheters are more easily visualized than on conventional images. Soft tissues of the chest wall and bony structures are also more clearly seen on the digital images. The authors' initial experience indicates that portable digital imaging of neonates is technically feasible and provides good-quality diagnostic images

  14. [Standard chest x-ray: from film to all digital!].

    Science.gov (United States)

    Ducou Le Pointe, H

    2000-04-01

    Conventional imaging will be totally replaced by digital imaging because of the lower exposure dosage and the clear advantages in terms of image processing, transfer and storage. Reports in the literature have demonstrated that a 2.5 pl/mm spatial resolution is satisfactory to detect a pneumothorax, a parenchymal nodule or a mediastinal mass. For most authors, this resolution is also sufficient to detect an interstitial syndrome. Digitalization by amplification or by particle detectors using wire chambers are not, for the time being, acceptable for interpretation of a simple chest image because of the low spatial resolution. The selenium cylinder is an adapted technology dedicated to chest imaging. Memory display screens can be used to digitalize the entire radiology unit. They are perfectly adapted for bedside imaging. Flat detectors are based on direct or indirect conversion of the X photons into an electrical signal during the clinical evaluation. The early results appear most promising.

  15. Nodule detection in digital chest radiography: Introduction to the radius chest trial

    International Nuclear Information System (INIS)

    Baath, M.; Haakansson, M.; Boerjesson, S.; Kheddache, S.; Grahn, A.; Ruschin, M.; Tingberg, A.; Mattsson, S.; Maansson, L. G.

    2005-01-01

    Most digital radiographic systems of today have wide latitude and are hence able to provide images with a small constraint on dose level. This opens up for an unprejudiced dose optimisation. However, in order to succeed in the optimisation task, good knowledge of the imaging and detection processes is needed. As a part of the European-wide research project 'unification of physical and clinical requirements for medical X-ray imaging - governed by the Radiological Imaging Unification Strategies (RADIUS) Group - a major image quality trial was conducted by members of the group. The RADIUS chest trial was focused on the detection of lung nodules in digital chest radiography with the aims of determining to what extent (1) the detection of a nodule is dependent on its location, (2) the system noise disturbs the detection of lung nodules, (3) the anatomical noise disturbs the detection of lung nodules and (4) the image background and anatomical background act as pure noise for the detection of lung nodules. The purpose of the present paper is to give an introduction to the trial and describe the framework and set-up of the investigation. (authors)

  16. Practical applications of digital tomosynthesis of the chest.

    Science.gov (United States)

    Galea, A; Durran, A; Adlan, T; Gay, D; Riordan, R; Dubbins, P; Williams, M P

    2014-04-01

    Digital tomosynthesis is a radiographic technique that generates a number of coronal raw images of a patient from a single pass of the x-ray tube. Tomosynthesis provides some of the tomographic benefits of computed tomography (CT), but at a much lower dose of radiation and cost when compared to CT. This review illustrates the range of practical applications of digital tomosynthesis of the chest. Copyright © 2013 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

  17. Evaluation of the image quality of chest CT scans: a phantom study

    Energy Technology Data Exchange (ETDEWEB)

    Martins N, P. I.; Prata M, A., E-mail: priscillainglid@gmail.com [Centro Federal de Educacao Tecnologica de Minas Gerais, Centro de Engenharia Biomedica, Av. Amazonas 5253, 30421-169 Nova Suica, Belo Horizonte, Minas Gerais (Brazil)

    2016-10-15

    Computed tomography (CT) is considered one of the most important methods of medical imaging employed nowadays, due to its non-invasiveness and the high quality of the images it is able to generate. However, the diagnostic radiation dose received by an individual over the year often exceeds the dose received on account of background radiation. Therefore, it is important to know and to control the dose distribution in the patient by varying the image acquisition parameters. The aim of this study is to evaluate the variation of the image quality of chest CT scans performed by two phantoms. In this paper, a cylindrical Polymethyl Methacrylate (PMMA) chest phantom was used and a second PMMA phantom has been developed with the same volume but an oblong shape, based on the actual dimensions of a male human thorax, in the axillary region. Ten-centimeter scans of the central area of each phantom were performed by a 16-channel Toshiba CT scanner, model Alexion. The scanning protocol employed was the radiology service protocol for chest scans. The noise survey was conducted within the image of the center slice, in five regions: one central and four peripheral areas close to the edge of the object (anterior, posterior, left and right). The recorded values showed that the oblong phantom, with a shape that is more similar to the actual human chest, has a considerably smaller noise, especially in the anterior, posterior and central regions. (Author)

  18. Evaluation of the image quality of chest CT scans: a phantom study

    International Nuclear Information System (INIS)

    Martins N, P. I.; Prata M, A.

    2016-10-01

    Computed tomography (CT) is considered one of the most important methods of medical imaging employed nowadays, due to its non-invasiveness and the high quality of the images it is able to generate. However, the diagnostic radiation dose received by an individual over the year often exceeds the dose received on account of background radiation. Therefore, it is important to know and to control the dose distribution in the patient by varying the image acquisition parameters. The aim of this study is to evaluate the variation of the image quality of chest CT scans performed by two phantoms. In this paper, a cylindrical Polymethyl Methacrylate (PMMA) chest phantom was used and a second PMMA phantom has been developed with the same volume but an oblong shape, based on the actual dimensions of a male human thorax, in the axillary region. Ten-centimeter scans of the central area of each phantom were performed by a 16-channel Toshiba CT scanner, model Alexion. The scanning protocol employed was the radiology service protocol for chest scans. The noise survey was conducted within the image of the center slice, in five regions: one central and four peripheral areas close to the edge of the object (anterior, posterior, left and right). The recorded values showed that the oblong phantom, with a shape that is more similar to the actual human chest, has a considerably smaller noise, especially in the anterior, posterior and central regions. (Author)

  19. Community-Acquired Pneumonia Visualized on CT Scans but Not Chest Radiographs: Pathogens, Severity, and Clinical Outcomes.

    Science.gov (United States)

    Upchurch, Cameron P; Grijalva, Carlos G; Wunderink, Richard G; Williams, Derek J; Waterer, Grant W; Anderson, Evan J; Zhu, Yuwei; Hart, Eric M; Carroll, Frank; Bramley, Anna M; Jain, Seema; Edwards, Kathryn M; Self, Wesley H

    2018-03-01

    The clinical significance of pneumonia visualized on CT scan in the setting of a normal chest radiograph is uncertain. In a multicenter prospective surveillance study of adults hospitalized with community-acquired pneumonia (CAP), we compared the presenting clinical features, pathogens present, and outcomes of patients with pneumonia visualized on a CT scan but not on a concurrent chest radiograph (CT-only pneumonia) and those with pneumonia visualized on a chest radiograph. All patients underwent chest radiography; the decision to obtain CT imaging was determined by the treating clinicians. Chest radiographs and CT images were interpreted by study-dedicated thoracic radiologists blinded to the clinical data. The study population included 2,251 adults with CAP; 2,185 patients (97%) had pneumonia visualized on chest radiography, whereas 66 patients (3%) had pneumonia visualized on CT scan but not on concurrent chest radiography. Overall, these patients with CT-only pneumonia had a clinical profile similar to those with pneumonia visualized on chest radiography, including comorbidities, vital signs, hospital length of stay, prevalence of viral (30% vs 26%) and bacterial (12% vs 14%) pathogens, ICU admission (23% vs 21%), use of mechanical ventilation (6% vs 5%), septic shock (5% vs 4%), and inhospital mortality (0 vs 2%). Adults hospitalized with CAP who had radiological evidence of pneumonia on CT scan but not on concurrent chest radiograph had pathogens, disease severity, and outcomes similar to patients who had signs of pneumonia on chest radiography. These findings support using the same management principles for patients with CT-only pneumonia and those with pneumonia seen on chest radiography. Copyright © 2017 American College of Chest Physicians. All rights reserved.

  20. Conversion to use of digital chest images for surveillance of coal workers' pneumoconiosis (black lung).

    Science.gov (United States)

    Levine, Betty A; Ingeholm, Mary Lou; Prior, Fred; Mun, Seong K; Freedman, Matthew; Weissman, David; Attfield, Michael; Wolfe, Anita; Petsonk, Edward

    2009-01-01

    To protect the health of active U.S. underground coal miners, the National Institute for Occupational Safety and Health (NIOSH) has a mandate to carry out surveillance for coal workers' pneumoconiosis, commonly known as Black Lung (PHS 2001). This is accomplished by reviewing chest x-ray films obtained from miners at approximately 5-year intervals in approved x-ray acquisition facilities around the country. Currently, digital chest images are not accepted. Because most chest x-rays are now obtained in digital format, NIOSH is redesigning the surveillance program to accept and manage digital x-rays. This paper highlights the functional and security requirements for a digital image management system for a surveillance program. It also identifies the operational differences between a digital imaging surveillance network and a clinical Picture Archiving Communication Systems (PACS) or teleradiology system.

  1. A study on the digital image transfer application mass chest X-ray system up-grade

    International Nuclear Information System (INIS)

    Kim, Sun Chil; Park, Jong Sam; Lee, Jon Il

    2003-01-01

    By converting movable indirect mass chest X-ray devices for vehicles into digital systems and upgrading it to share information with the hospital's medical image information system, excellencies have been confirmed as a result of installing and running this type of system and are listed hereinafter. Upgrading analog systems, such as indirect mass chest X-ray devices dependent on printed film, to digital systems allows them to be run and managed much more efficiently, contributing to the increase in the stability and the efficiency of the system. Unlike existing images, communication based on DICOM standards allow images to be compatible with the hospital's outer and inner network PACS systems, extending the scope of the radiation departments information system. Assuming chest-exclusive indirect mass chest X-rays, a linked development of CAD (Computer Aided Diagnosis, Detector) becomes possible. By applying wireless Internet, Web-PACS for movable indirect mass chest X-ray devices for vehicles will become possible. Research in these fields must continue and if the superior image quality and convenience of digital systems are confirmed, I believe that the conversion of systems still dependent on analog images to modernized digital systems is a must

  2. Segmentation of ribs in digital chest radiographs

    Science.gov (United States)

    Cong, Lin; Guo, Wei; Li, Qiang

    2016-03-01

    Ribs and clavicles in posterior-anterior (PA) digital chest radiographs often overlap with lung abnormalities such as nodules, and cause missing of these abnormalities, it is therefore necessary to remove or reduce the ribs in chest radiographs. The purpose of this study was to develop a fully automated algorithm to segment ribs within lung area in digital radiography (DR) for removal of the ribs. The rib segmentation algorithm consists of three steps. Firstly, a radiograph was pre-processed for contrast adjustment and noise removal; second, generalized Hough transform was employed to localize the lower boundary of the ribs. In the third step, a novel bilateral dynamic programming algorithm was used to accurately segment the upper and lower boundaries of ribs simultaneously. The width of the ribs and the smoothness of the rib boundaries were incorporated in the cost function of the bilateral dynamic programming for obtaining consistent results for the upper and lower boundaries. Our database consisted of 93 DR images, including, respectively, 23 and 70 images acquired with a DR system from Shanghai United-Imaging Healthcare Co. and from GE Healthcare Co. The rib localization algorithm achieved a sensitivity of 98.2% with 0.1 false positives per image. The accuracy of the detected ribs was further evaluated subjectively in 3 levels: "1", good; "2", acceptable; "3", poor. The percentages of good, acceptable, and poor segmentation results were 91.1%, 7.2%, and 1.7%, respectively. Our algorithm can obtain good segmentation results for ribs in chest radiography and would be useful for rib reduction in our future study.

  3. Digital luminescent radiography: A substitute for conventional chest radiography?

    International Nuclear Information System (INIS)

    Neufang, K.F.R.; Krug, B.; Lorenz, R.; Steinbrich, W.

    1990-01-01

    The image quality of digital luminescent radiography (DLR) is sufficient for routine biplane chest radiography and for follow-up studies of heart size, pulmonary congestion, coin lesions, infiltrations, atelectasis, pleural effusions, and mediastinal and hilar lymph node enlargement. Chest radiography in the intensive care unit may in most cases be performed using the DLR technique. there is no need for repeat shots because of incorrect exposure, and the position of catheters, tubes, pacemakers, drains and artificial heart valves, the mediastinum, and the retrocardiac areas of the left lung are more confidently assessed on the edge-enhanced DLR films than on conventional films. Nevertheless, DLR is somewhat inferior to conventional film-screen radiography of the chest as it can demonstrate or rule out subtle pulmonary interstitial disease less confidently. There is no reduction of radiation exposure of the chest in DLR compared with modern film-screen systems. As a consequence, DLR is presently not in a position to replace traditional film-screen radiography of the chest completely. (orig.) [de

  4. Three-dimensional hard and soft tissue imaging of the human cochlea by scanning laser optical tomography (SLOT.

    Directory of Open Access Journals (Sweden)

    Nadine Tinne

    Full Text Available The present study focuses on the application of scanning laser optical tomography (SLOT for visualization of anatomical structures inside the human cochlea ex vivo. SLOT is a laser-based highly efficient microscopy technique which allows for tomographic imaging of the internal structure of transparent specimens. Thus, in the field of otology this technique is best convenient for an ex vivo study of the inner ear anatomy. For this purpose, the preparation before imaging comprises decalcification, dehydration as well as optical clearing of the cochlea samples in toto. Here, we demonstrate results of SLOT imaging visualizing hard and soft tissue structures with an optical resolution of down to 15 μm using extinction and autofluorescence as contrast mechanisms. Furthermore, the internal structure can be analyzed nondestructively and quantitatively in detail by sectioning of the three-dimensional datasets. The method of X-ray Micro Computed Tomography (μCT has been previously applied to explanted cochlea and is solely based on absorption contrast. An advantage of SLOT is that it uses visible light for image formation and thus provides a variety of contrast mechanisms known from other light microscopy techniques, such as fluorescence or scattering. We show that SLOT data is consistent with μCT anatomical data and provides additional information by using fluorescence. We demonstrate that SLOT is applicable for cochlea with metallic cochlear implants (CI that would lead to significant artifacts in μCT imaging. In conclusion, the present study demonstrates the capability of SLOT for resolution visualization of cleared human cochleae ex vivo using multiple contrast mechanisms and lays the foundation for a broad variety of additional studies.

  5. Early Chest Computed Tomography Scan to Assist Diagnosis and Guide Treatment Decision for Suspected Community-acquired Pneumonia.

    Science.gov (United States)

    Claessens, Yann-Erick; Debray, Marie-Pierre; Tubach, Florence; Brun, Anne-Laure; Rammaert, Blandine; Hausfater, Pierre; Naccache, Jean-Marc; Ray, Patrick; Choquet, Christophe; Carette, Marie-France; Mayaud, Charles; Leport, Catherine; Duval, Xavier

    2015-10-15

    Clinical decision making relative to community-acquired pneumonia (CAP) diagnosis is difficult. Chest radiograph is key in establishing parenchymal lung involvement. However, radiologic performance may lead to misdiagnosis, rendering questionable the use of chest computed tomography (CT) scan in patients with clinically suspected CAP. To assess whether early multidetector chest CT scan affects diagnosis and management of patients visiting the emergency department with suspected CAP. A total of 319 prospectively enrolled patients with clinically suspected CAP underwent multidetector chest CT scan within 4 hours. CAP diagnosis probability (definite, probable, possible, or excluded) and therapeutic plans (antibiotic initiation/discontinuation, hospitalization/discharge) were established by emergency physicians before and after CT scan results. The adjudication committee established the final CAP classification on Day 28. Chest radiograph revealed a parenchymal infiltrate in 188 patients. CAP was initially classified as definite in 143 patients (44.8%), probable or possible in 172 (53.8%), and excluded in 4 (1.2%). CT scan revealed a parenchymal infiltrate in 40 (33%) of the patients without infiltrate on chest radiograph and excluded CAP in 56 (29.8%) of the 188 with parenchymal infiltrate on radiograph. CT scan modified classification in 187 (58.6%; 95% confidence interval, 53.2-64.0), leading to 50.8% definite CAP and 28.8% excluded CAP, and 80% of modifications were in accordance with adjudication committee classification. Because of CT scan, antibiotics were initiated in 51 (16%) and discontinued in 29 (9%), and hospitalization was decided in 22 and discharge in 23. In CAP-suspected patients visiting the emergency unit, early CT scan findings complementary to chest radiograph markedly affect both diagnosis and clinical management. Clinical trial registered with www.clinicaltrials.gov (NCT 01574066).

  6. Impact of matrix size on observer performance in digital chest radiography

    International Nuclear Information System (INIS)

    Koelblinger, C.

    2003-02-01

    This thesis compared the observer performance in the detection of abnormalities on 2k matrix(0,2 mm pixel size) and 4k matrix (0,1 mm pixel size) digital chest radiographs. Eighty five patients who underwent CT of the thorax were prospectively reccruited into the study. A chest x-ray of each patient was acquired in 2k and 4k format and four readers analyzed the images by different criterias. On the one hand a ROC analysis was performed with the CT data as goldstandard. On the other hand each reader had to rate different anatomical structures in a direct comparison of the pictures. The results of the ratings did not show any significant difference between the 2k and the 4k format. The conclusion or this study is that the use of a 4k instead of a 2k matrix in digital chest radiography does not yield to an improved observer performance. (author)

  7. Digital radiography of the chest in pediatric patients

    International Nuclear Information System (INIS)

    Puig, S.

    2003-01-01

    The hopes placed in digital radiography have been fulfilled only partly in pediatric radiology. Specifically, the option of gaining reduced radiation exposure in combination with a similar or even improved image quality was hard to realize. The only portable digital system available for a long time were storage phosphors which were disadvantaged by an extremely limited dose-quantum-efficiency (DQE) in comparison to digital flat panel detectors. New developments and the introduction of the dual-reading system led to image qualities comparable to film-screen-systems with high resolution and achievable without dose increase, sometimes even with dose reduction. A study using an animal model suggests that these systems can even be used in preterm infants with very low birth weights. A new portable flat panel detector by Canon may improve digital chest radiography in pediatric patients. (orig.) [de

  8. SU-D-204-06: Dose and Image Quality Evaluation of a Low-Dose Slot-Scanning X-Ray System for Pediatric Orthopedic Studies

    Energy Technology Data Exchange (ETDEWEB)

    Liu, Z; Hoerner, M; Lamoureux, R; Rill, L; Arreola, M [Univ Florida, Jacksonville Beach, FL (United States)

    2015-06-15

    Purpose: Children in early teens with scoliosis require repeated radiographic exams over a number of years. The EOS (EOS imaging S.A., Paris, France) is a novel low-dose slot-scanning digital radiographic system designed to produce full-spine images of a free-standing patient. The radiation dose and image quality characteristics of the EOS were evaluated relative to those of a Computed Radiography (CR) system for scoliosis imaging. Methods: For dose evaluation, a full-torso anthropomorphic phantom was scanned five times using the default standard clinical protocols for both the EOS and a CR system, which include both posteroanterior and lateral full-spine views. Optically stimulated luminescent dosimeters (OSLDs), also known as nanoDots™ (Landauer, Inc., Glenwood, IL), were placed on the phantom’s surface to measure entrance skin dose. To assess image quality, MTF curves were generated from sampling the noise levels within the high-contrast regions of a line-pair phantom. Vertical and horizontal distortions were measured for the square line-pair phantom with the EOS system to evaluate the effects of geometric magnification and misalignment with the indicated imaging plane. Results: The entrance skin dose was measured to be 0.4 to 1.1 mGy for the EOS, and 0.7 to 3.6 mGy for the CR study. MTF comparison shows that CR greatly outperforms the EOS, despite both systems having a limiting resolution at 1.8 line-pairs per mm. Vertical distortion was unaffected by phantom positioning, because of the EOS slot-scanning geometry. Horizontal distortion increased linearly with miscentering distance. Conclusion: The EOS system resulted in approximately 70% lower radiation dose than CR for full-spine images. Image quality was found to be inferior to CR. Further investigation is required to see if EOS system is an acceptable modality for performing clinically diagnostic scoliosis examinations.

  9. SU-D-204-06: Dose and Image Quality Evaluation of a Low-Dose Slot-Scanning X-Ray System for Pediatric Orthopedic Studies

    International Nuclear Information System (INIS)

    Liu, Z; Hoerner, M; Lamoureux, R; Rill, L; Arreola, M

    2015-01-01

    Purpose: Children in early teens with scoliosis require repeated radiographic exams over a number of years. The EOS (EOS imaging S.A., Paris, France) is a novel low-dose slot-scanning digital radiographic system designed to produce full-spine images of a free-standing patient. The radiation dose and image quality characteristics of the EOS were evaluated relative to those of a Computed Radiography (CR) system for scoliosis imaging. Methods: For dose evaluation, a full-torso anthropomorphic phantom was scanned five times using the default standard clinical protocols for both the EOS and a CR system, which include both posteroanterior and lateral full-spine views. Optically stimulated luminescent dosimeters (OSLDs), also known as nanoDots™ (Landauer, Inc., Glenwood, IL), were placed on the phantom’s surface to measure entrance skin dose. To assess image quality, MTF curves were generated from sampling the noise levels within the high-contrast regions of a line-pair phantom. Vertical and horizontal distortions were measured for the square line-pair phantom with the EOS system to evaluate the effects of geometric magnification and misalignment with the indicated imaging plane. Results: The entrance skin dose was measured to be 0.4 to 1.1 mGy for the EOS, and 0.7 to 3.6 mGy for the CR study. MTF comparison shows that CR greatly outperforms the EOS, despite both systems having a limiting resolution at 1.8 line-pairs per mm. Vertical distortion was unaffected by phantom positioning, because of the EOS slot-scanning geometry. Horizontal distortion increased linearly with miscentering distance. Conclusion: The EOS system resulted in approximately 70% lower radiation dose than CR for full-spine images. Image quality was found to be inferior to CR. Further investigation is required to see if EOS system is an acceptable modality for performing clinically diagnostic scoliosis examinations

  10. Radiochromic film use to record dose profile variations in chest CT scan; Utilizacao do filme radiocromico para registro da variacao do perfil de dose em varredura de TC de torax

    Energy Technology Data Exchange (ETDEWEB)

    Mourao, Arnaldo P.; Santana, Priscila C., E-mail: aprata@des.cefetmg.br [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear; Silva, Teogenes A. da; Alonso, Thessa C. [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2015-12-15

    The CT scans generate higher doses in patients than those caused by other types of diagnostic radiology using the attenuation of X-ray beams. Technological advances have increased the clinical applications of computed tomography (CT) and consequently the demand for these exams. Dose reduction strategies are difficult to implement because of a lack of proper guidance on the CT scanning. However, CT offers the possibility of adjusting the image acquisition parameters according to the patient physical profile and the diagnostic application for which it is intended to scan. Knowledge of the dose distribution is important when thinking about varying the acquisition parameters to reduce the dose. In this work was used a PMMA chest phantom to observe the dose deposition in 5 areas. Radiochromic film strips were placed inside the chest phantom, in peripheral and in the central region. The phantom was placed in the scanner isocenter and it was performed the irradiation of its central region using a chest acquisition protocol. After the phantom CT scan the radiochromic film strips were processed for obtaining digital images. Digital images were reworked to obtain the dose variation profiles for each position. The results showed a wide variation in absorbed dose by the phantom, either within a same position along the length of the film strip, as in the comparison among the five regions which the strips are placed. In this second case the dose variations were even greater. (author)

  11. A dual-view digital tomosynthesis imaging technique for improved chest imaging

    Energy Technology Data Exchange (ETDEWEB)

    Zhong, Yuncheng; Lai, Chao-Jen; Wang, Tianpeng; Shaw, Chris C., E-mail: cshaw@mdanderson.org [Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77054 (United States)

    2015-09-15

    Purpose: Digital tomosynthesis (DTS) has been shown to be useful for reducing the overlapping of abnormalities with anatomical structures at various depth levels along the posterior–anterior (PA) direction in chest radiography. However, DTS provides crude three-dimensional (3D) images that have poor resolution in the lateral view and can only be displayed with reasonable quality in the PA view. Furthermore, the spillover of high-contrast objects from off-fulcrum planes generates artifacts that may impede the diagnostic use of the DTS images. In this paper, the authors describe and demonstrate the use of a dual-view DTS technique to improve the accuracy of the reconstructed volume image data for more accurate rendition of the anatomy and slice images with improved resolution and reduced artifacts, thus allowing the 3D image data to be viewed in views other than the PA view. Methods: With the dual-view DTS technique, limited angle scans are performed and projection images are acquired in two orthogonal views: PA and lateral. The dual-view projection data are used together to reconstruct 3D images using the maximum likelihood expectation maximization iterative algorithm. In this study, projection images were simulated or experimentally acquired over 360° using the scanning geometry for cone beam computed tomography (CBCT). While all projections were used to reconstruct CBCT images, selected projections were extracted and used to reconstruct single- and dual-view DTS images for comparison with the CBCT images. For realistic demonstration and comparison, a digital chest phantom derived from clinical CT images was used for the simulation study. An anthropomorphic chest phantom was imaged for the experimental study. The resultant dual-view DTS images were visually compared with the single-view DTS images and CBCT images for the presence of image artifacts and accuracy of CT numbers and anatomy and quantitatively compared with root-mean-square-deviation (RMSD) values

  12. Radiology of pulmonary disease. Chest radiography, computed tomography, and gallium scanning

    International Nuclear Information System (INIS)

    Golden, J.A.; Sollitto, R.A.

    1988-01-01

    A review of the radiologic manifestations of AIDS pulmonary diseases, with an emphasis on the utility of gallium scanning in the context of the normal or equivocal chest x-ray, is presented.99 references

  13. [Optimization of digital chest radiography image post-processing in diagnosis of pneumoconiosis].

    Science.gov (United States)

    Sheng, Bing-yong; Mao, Ling; Zhou, Shao-wei; Shi, Jin

    2013-11-01

    To establish the optimal image post-processing parameters for digital chest radiography as preliminary research for introducing digital radiography (DR) to pneumoconiosis diagnosis in China. A total of 204 pneumoconiosis patients and 31 dust-exposed workers were enrolled as the subjects in this research. Film-screen radiography (FSR) and DR images were taken for all subjects. DR films were printed after raw images were processed and parameters were altered using DR workstation (GE Healthcare, U.S.A.). Image gradations, lung textures, and the imaging of thoracic vertebra were evaluated by pneumoconiosis experts, and the optimal post-processing parameters were selected. Optical density was measured for both DR films and FSR films. For the DR machine used in this research, the contrast adjustment (CA) and brightness adjustment (BA) were the main parameters that determine the brightness and gray levels of images. The optimal ranges for CA and BA were 115%∼120% and 160%∼165%, respectively. The quality of DR chest films would be optimized when tissue contrast was adjusted to a maximum of 0.15, edge to a minimum of 1, and both noise reduction and tissue equalization to0.The failure rate of chest DR (0.4%) was significantly lower than that of chest FSR (17%) (P image post-processing on DR machine purchased from GE Healthcare, the DR chest films can meet all requirements for the quality of chest X-ray films in the Chinese diagnostic criteria for pneumoconiosis.

  14. Computerized detection of pneumothorax on digital chest radiographs

    International Nuclear Information System (INIS)

    Sanada, S.; Doi, K.; MacMahon, H.; Montner, S.M.

    1990-01-01

    This paper reports on neumothoraces that are clinically important abnormalities that usually appear as a subtle, fine line pattern on chest radiographs. We are developing a computer vision system for automated detection of pneumothorax to aid radiologists diagnosis. Chest images were digitized with a 0.175-mm pixel size, yielding a 2,000 x 2,430 matrix size, and 10 bits of gray scale. After indentification of the lung regions, an edge detection filter was employed in the apical areas to enhance a pneumothorax pattern. Ribs were detected with a technique based on statistical analysis of edge gradients and their orientations. Points located on a curved line suggestive of a pneumothorax in this enhanced image were detected with a Hough transform

  15. Development of a fully automated adaptive unsharp masking technique in digital chest radiograph

    International Nuclear Information System (INIS)

    Abe, Katsumi; Katsuragawa, Shigehiko; Sasaki, Yasuo

    1991-01-01

    We are developing a fully automated adaptive unsharp masking technique with various parameters depending on regional image features of a digital chest radiograph. A chest radiograph includes various regions such as lung fields, retrocardiac area and spine in which their texture patterns and optical densities are extremely different. Therefore, it is necessary to enhance image contrast of each region by each optimum parameter. First, we investigated optimum weighting factors and mask sizes of unsharp masking technique in a digital chest radiograph. Then, a chest radiograph is automatically divided into three segments, one for the lung field, one for the retrocardiac area, and one for the spine, by using histogram analysis of pixel values. Finally, high frequency components of the lung field and retrocardiac area are selectively enhanced with a small mask size and mild weighting factors which are previously determined as optimum parameters. In addition, low frequency components of the spine are enhanced with a large mask size and adequate weighting factors. This processed image shows excellent depiction of the lung field, retrocardiac area and spine simultaneously with optimum contrast. Our image processing technique may be useful for diagnosis of chest radiographs. (author)

  16. Radiological study of the chest using the digital method (FCR 101)

    International Nuclear Information System (INIS)

    Ferri, G.; Battolla, L.; Palla, L.; D'Alessandro, F.; Torri, T.; Tagliagambe, A.

    1991-01-01

    A computed radiography system with solid state detectors (FCR 101) was employed to evaluate the quality of digital images, with optimal and progressively lowered doses, in the study of the chest. Routine chest radiographs of in patients with no pathologic conditions of the chest were studied. The patients were divided into 3 groups: each patient in group A underwent two radiographs, one at 100% andthe other at 50% exposure; patients in group B were imaged with 100% and 25% doses, and those in group C received 50% and 25% exposure. Several parameters were employed for image evaluation, and a value was given to each of them (1=good, 2=sufficient, 3=insufficient). Upon comparison of the average values obtained, we observed that the digital technique allowed a most accurate and well-defined representation of the examined parameters at 100% exposure dose and that a 50% reduction did not decrease the informative content of the least. A 25% reduction, however, determined a high degree of background noise and a subsequent, though slight, loss of information

  17. Effect of data compression on diagnostic accuracy in digital hand and chest radiography

    Science.gov (United States)

    Sayre, James W.; Aberle, Denise R.; Boechat, Maria I.; Hall, Theodore R.; Huang, H. K.; Ho, Bruce K. T.; Kashfian, Payam; Rahbar, Guita

    1992-05-01

    Image compression is essential to handle a large volume of digital images including CT, MR, CR, and digitized films in a digital radiology operation. The full-frame bit allocation using the cosine transform technique developed during the last few years has been proven to be an excellent irreversible image compression method. This paper describes the effect of using the hardware compression module on diagnostic accuracy in hand radiographs with subperiosteal resorption and chest radiographs with interstitial disease. Receiver operating characteristic analysis using 71 hand radiographs and 52 chest radiographs with five observers each demonstrates that there is no statistical significant difference in diagnostic accuracy between the original films and the compressed images with a compression ratio as high as 20:1.

  18. Aortic valve calcification - a commonly observed but frequently ignored finding during CT scanning of the chest.

    Science.gov (United States)

    Raju, Prashanth; Sallomi, David; George, Bindu; Patel, Hitesh; Patel, Nikhil; Lloyd, Guy

    2012-06-01

    To describe the frequency and severity of Aortic valve calcification (AVC) in an unselected cohort of patients undergoing chest CT scanning and to assess the frequency with which AVC was being reported in the radiology reports. Consecutive CT scan images of the chest and the radiological reports (December 2009 to May 2010) were reviewed at the district general hospital (DGH). AVC on CT scan was visually graded on a scale ranging from 0 to IV (0 = no calcification, IV = severe calcification). Total of 416 (232 male; 184 female) CT chest scans [Contrast enhanced 302 (72%), unenhanced 114 (28%)] were reviewed. Mean age was 70.55 ± 11.48 years. AVC in CT scans was identified in 95 of the 416 patients (22.83%). AVC classification was as follows: Grade I: 60 (63.15%), Grade II: 22 (23.15%), Grade III: 9 (9.47%), Grade IV: 4 (4.21%). Only one CT report mentioned AVC. Only 31 of 95 AVC had Transthoracic echocardiogram (TTE). The interval time between CT scan and TTE was variable.   Aortic valve calcification in CT chest scans is a common finding and studies have shown that it is strongly related to the presence and severity of aortic valve disease. As CT scans are considered as a valuable additional screening tool for detection of aortic stenosis, AVC should always be commented upon in the radiology reports. Furthermore, patients with at least Grade III and IV AVC should be sent for TTE. © 2012 Blackwell Publishing Ltd.

  19. Evaluation of Image Quality in Low Tube-Voltage Chest CT Scan

    International Nuclear Information System (INIS)

    Kim, Hyun Ju; Cho, Jae Hwan; Park, Cheol Soo

    2010-01-01

    The patients who visited this department for pulmonary disease and need CT scans for Follow-up to observe change of CT value, evaluation of image quality and decrease of radiation dose as change of kVp. Subjects were the patients of 20 persons visited this department for pulmonary disease and Somatom Sensation 16(Semens, Enlarge, Germany) was used. Measurement of CT value as change of kVp was done by setting up ROI diameter of 1cm at the height of thyroid, aortic arch, right pulmonary artery in arterial phase image using 100 kVp, measuring 3 times, and recorded the average. CT value of phantom was measured by scanning phantoms which means contrast media diluted by normal saline by various ratio with tube voltage of 80 kVp, 100 kVp, 120 kVp, 140 kVp and recorded the average of 3 CT values of center of phantom image. In analysing radiation dose, CTDIVOL values of the latest arterial phase image of 120 kVp and as this research set that of 100 kVp were analyzed comparatively. 2 observers graded quality of chest images by 5 degrees (Unacceptable, Suboptimal, Adequate, Good, Excellent). CT value of chest image increased at 100 kVp by 14.06%∼27.26% in each ROI than 120 kVp. CT value of phantom increased as tube voltage lowered at various concentration of contrast media. CTDIVOL decreased at 100 kVp(5.00 mGy) by 36% than 120 kVp(7.80 mGy) in radiation dose analysis. here were 0 Unacceptable, 1 Suboptimal, 3 Adequate, 10 Good, 6 Excellent in totally 20 persons. Chest CT scanning with low kilo-voltage for patients who need CT scan repeatedly can bring images valuable for diagnose, and decrease radiation dose against patients

  20. Quantitative evaluation of changes in pleural effusion using digital images of the chest

    International Nuclear Information System (INIS)

    Nishiike, Shigeaki; Fujimura, Ichiro; Sagara, Kenji; Sakashita, Keiji

    2000-01-01

    Exposure data recognition (EDR) is a function of FCR that automatically determines the optimal density of images. This function makes it difficult to evaluate changes in pleural effusion on sequential chest CR. We developed a method to quantify the amount of pleural effusion using the digital value and S value of chest CR. We evaluated the efficacy of the method to quantify the amount of pleural effusion using these values of chest CR in clinical settings. The study period was from December 1996 to January 1999. Study 1: Chest CR was carried out before and after pleural cavity drainage in 70 cases. The volume calculated with the method using the digital value and S value of CR [x(ml)] was compared with the drained volume [y1] on both sides. Study 2: Change in pleural effusion was estimated with a method using CT (y2) and with the method using the digital value and S value of CR (x). The relationship of these two data was investigated. Study 1: The regression equations between x and y1 were y1=1.05x+21.1 (r 2 =0.89) on the right side and y1=0.95x+69.7 (r 2 =0.68) on the left side. Study 2: The regression equations between x and y2 were y2=0.75x-1.88 (r 2 =0.70) on the right side and y2=0.87x-1.27 (r 2 =0.65) on the left side. The method using the digital value and S value of CR was effective for quantifying changes in pleural effusion. (author)

  1. Development of a prototype chest digital tomosynthesis (CDT) R/F system with fast image reconstruction using graphics processing unit (GPU) programming

    Energy Technology Data Exchange (ETDEWEB)

    Choi, Sunghoon, E-mail: choi.sh@yonsei.ac.kr [Department of Radiological Science, College of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju, Gangwon-do 220-710 (Korea, Republic of); Lee, Seungwan [Department of Radiological Science, College of Medical Science, Konyang University, 158 Gwanjeodong-ro, Daejeon, 308-812 (Korea, Republic of); Lee, Haenghwa [Department of Radiological Science, College of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju, Gangwon-do 220-710 (Korea, Republic of); Lee, Donghoon; Choi, Seungyeon [Department of Radiation Convergence Engineering, College of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju, Gangwon-do 220-710 (Korea, Republic of); Shin, Jungwook [LISTEM Corporation, 94 Donghwagongdan-ro, Munmak-eup, Wonju (Korea, Republic of); Seo, Chang-Woo [Department of Radiological Science, College of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju, Gangwon-do 220-710 (Korea, Republic of); Kim, Hee-Joung, E-mail: hjk1@yonsei.ac.kr [Department of Radiological Science, College of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju, Gangwon-do 220-710 (Korea, Republic of); Department of Radiation Convergence Engineering, College of Health Science, Yonsei University, 1 Yonseidae-gil, Wonju, Gangwon-do 220-710 (Korea, Republic of)

    2017-03-11

    Digital tomosynthesis offers the advantage of low radiation doses compared to conventional computed tomography (CT) by utilizing small numbers of projections (~80) acquired over a limited angular range. It produces 3D volumetric data, although there are artifacts due to incomplete sampling. Based upon these characteristics, we developed a prototype digital tomosynthesis R/F system for applications in chest imaging. Our prototype chest digital tomosynthesis (CDT) R/F system contains an X-ray tube with high power R/F pulse generator, flat-panel detector, R/F table, electromechanical radiographic subsystems including a precise motor controller, and a reconstruction server. For image reconstruction, users select between analytic and iterative reconstruction methods. Our reconstructed images of Catphan700 and LUNGMAN phantoms clearly and rapidly described the internal structures of phantoms using graphics processing unit (GPU) programming. Contrast-to-noise ratio (CNR) values of the CTP682 module of Catphan700 were higher in images using a simultaneous algebraic reconstruction technique (SART) than in those using filtered back-projection (FBP) for all materials by factors of 2.60, 3.78, 5.50, 2.30, 3.70, and 2.52 for air, lung foam, low density polyethylene (LDPE), Delrin{sup ®} (acetal homopolymer resin), bone 50% (hydroxyapatite), and Teflon, respectively. Total elapsed times for producing 3D volume were 2.92 s and 86.29 s on average for FBP and SART (20 iterations), respectively. The times required for reconstruction were clinically feasible. Moreover, the total radiation dose from our system (5.68 mGy) was lower than that of conventional chest CT scan. Consequently, our prototype tomosynthesis R/F system represents an important advance in digital tomosynthesis applications.

  2. Protocol optimization in chest CT scans of child

    Energy Technology Data Exchange (ETDEWEB)

    Abrao L, L. T.; Amaral de O, F.; Prata M, A. [Biomedical Engineering Center, Centro Federal de Educacao Tecnologica de Minas Gerais, 30421-169, Belo Horizonte, Minas Gerais (Brazil); Bustos F, M., E-mail: luanaabrao@gmail.com [Universidad Federal de Minas Gerais, Department of Nuclear Engineering, Av. Pres. Antonio Carlos 6627, Pampulha, 31270-901 Belo Horizonte, Minas Gerais (Brazil)

    2017-10-15

    The dissemination of Computed Tomography (CT), a radiodiagnostic technique, has significant increase in the patient dose. In the last years, this technique has shown a high growth due to clinical cases of medical emergencies, neoplasm and pediatric traumas. Dose measurement is important to correlate with the deleterious effects of radiation on the organism and radiation future effects is related with stochastic risks due to tissue radiosensitivity, allied to the life expectancy of the child. In this work, a cylindrical phantom, representing an adult chest made of polymethylmethacrylate (PMMA), was used and a new born chest phantom with a shape oblong was developed based on the dimensions of a typical newborn. In a Ge CT scanner, Discovery model, with 64 channels, the central slice of the phantoms were irradiated successively in order to obtain dose measurements using an ionizing pencil camera. Based in the measurements, dose index was calculated (CTDI{sub vol}). The radiological service chest protocol using a voltage of 120 kV was used for scanning 10 cm of the central area of the adult and newborn phantom, in helical mode. An acquisition of images was performed using this radiological service chest protocol to compare with the protocol optimized. In the newborn phantom was also used protocols optimized using a voltage of 120 and 80 kV. The voltage of 80 kV has the lowest dose index for the pediatric object phantom. This work allowed the comparison between absorbed dose variations by the pediatric phantom changing the X-ray tube supply voltage. This dose variation has shown how important is specific protocols for children. (Author)

  3. Protocol optimization in chest CT scans of child

    International Nuclear Information System (INIS)

    Abrao L, L. T.; Amaral de O, F.; Prata M, A.; Bustos F, M.

    2017-10-01

    The dissemination of Computed Tomography (CT), a radiodiagnostic technique, has significant increase in the patient dose. In the last years, this technique has shown a high growth due to clinical cases of medical emergencies, neoplasm and pediatric traumas. Dose measurement is important to correlate with the deleterious effects of radiation on the organism and radiation future effects is related with stochastic risks due to tissue radiosensitivity, allied to the life expectancy of the child. In this work, a cylindrical phantom, representing an adult chest made of polymethylmethacrylate (PMMA), was used and a new born chest phantom with a shape oblong was developed based on the dimensions of a typical newborn. In a Ge CT scanner, Discovery model, with 64 channels, the central slice of the phantoms were irradiated successively in order to obtain dose measurements using an ionizing pencil camera. Based in the measurements, dose index was calculated (CTDI vol ). The radiological service chest protocol using a voltage of 120 kV was used for scanning 10 cm of the central area of the adult and newborn phantom, in helical mode. An acquisition of images was performed using this radiological service chest protocol to compare with the protocol optimized. In the newborn phantom was also used protocols optimized using a voltage of 120 and 80 kV. The voltage of 80 kV has the lowest dose index for the pediatric object phantom. This work allowed the comparison between absorbed dose variations by the pediatric phantom changing the X-ray tube supply voltage. This dose variation has shown how important is specific protocols for children. (Author)

  4. Digital chest radiography

    DEFF Research Database (Denmark)

    Debess, Jeanne Elisabeth; Vejle-Sørensen, Jens Kristian; Thomsen, Henrik

    ,3 mAs and SID SID of 180 centimetres using a phantom and lithium fluoride thermo luminescence dosimeter (TLD). Dose to risk organs mamma, thyroid and colon are measured at different collimations with one-centimetre steps. TLD results are used to estimate dose reduction for different collimations...... at the conference. Conclusion: Collimation improvement in basic chest radiography can reduce the radiation to female patients at chest x-ray examinations....

  5. Anatomical decomposition in dual energy chest digital tomosynthesis

    Science.gov (United States)

    Lee, Donghoon; Kim, Ye-seul; Choi, Sunghoon; Lee, Haenghwa; Choi, Seungyeon; Kim, Hee-Joung

    2016-03-01

    Lung cancer is the leading cause of cancer death worldwide and the early diagnosis of lung cancer has recently become more important. For early screening lung cancer, computed tomography (CT) has been used as a gold standard for early diagnosis of lung cancer [1]. The major advantage of CT is that it is not susceptible to the problem of misdiagnosis caused by anatomical overlapping while CT has extremely high radiation dose and cost compared to chest radiography. Chest digital tomosynthesis (CDT) is a recently introduced new modality for lung cancer screening with relatively low radiation dose compared to CT [2] and also showing high sensitivity and specificity to prevent anatomical overlapping occurred in chest radiography. Dual energy material decomposition method has been proposed for better detection of pulmonary nodules as means of reducing the anatomical noise [3]. In this study, possibility of material decomposition in CDT was tested by simulation study and actual experiment using prototype CDT. Furthermore organ absorbed dose and effective dose were compared with single energy CDT. The Gate v6 (Geant4 application for tomographic emission), and TASMIP (Tungsten anode spectral model using the interpolating polynomial) code were used for simulation study and simulated cylinder shape phantom consisted of 4 inner beads which were filled with spine, rib, muscle and lung equivalent materials. The patient dose was estimated by PCXMC 1.5 Monte Carlo simulation tool [4]. The tomosynthesis scan was performed with a linear movement and 21 projection images were obtained over 30 degree of angular range with 1.5° degree of angular interval. The proto type CDT system has same geometry with simulation study and composed of E7869X (Toshiba, Japan) x-ray tube and FDX3543RPW (Toshiba, Japan) detector. The result images showed that reconstructed with dual energy clearly visualize lung filed by removing unnecessary bony structure. Furthermore, dual energy CDT could enhance

  6. Future approaches to digital systems

    International Nuclear Information System (INIS)

    Woodrough, R.E.; Beckmann, E.; Jenkins, D.; Belcher, D.

    1984-01-01

    The introduction of the first digital imaging systems has led to the concept of a totally digital imaging department - the 'filmless department'. This concept and its associated network system for digital departments are briefly considered. A transducer system of scanned projection radiography with solid state detectors used in digital chest imaging is described. The principles of this transducer system are applicable to other forms of digital radiography. (U.K.)

  7. Flat panel detectors - closing the (digital) gap in chest and skeletal radiology

    International Nuclear Information System (INIS)

    Reiff, Kurt J.

    1999-01-01

    In the radiological department today the majority of all X-ray procedures on chest and skeletal radiography is performed with classical film-screen-systems. Using digital luminescence radiography (DLR or CR, which stands for Computed Radiography) as a technique has shown a way to replace this 100-year-old procedure of doing general radiography work by acquiring the X-rays digitally via phosphor screens, but this approach has faced criticism from lots of radiologists world wide and therefore has not been widely accepted except in the intensive care environment. A new technology is now rising based on the use of so called flat panel X-ray (FD) detectors. Semi-conducting material detects the X-rays in digital form directly and creates an instantaneous image for display, distribution and diagnosis. This ability combined with a large field of view and -- compared to existing methods -- excellent detective quantum efficiency represents a revolutionary step for chest and skeletal radiography and will put basic X-ray-work back into the focus of radiological solutions. This paper will explain the basic technology of flat panel detectors, possible system solutions based on this new technology, aspects of the user interface influencing the system utilization and versatility as well as the possibility to redefine the patient examination process for chest and skeletal radiography. Furthermore the author discusses limitations for the first released systems, upgrades for the installed base and possible scenarios for the future. e.g. fluoroscopy or angiography application

  8. Scanning negatives and slides digitizing your photographic archive

    CERN Document Server

    Steinhoff, Sascha

    2009-01-01

    Many photographers have either moved into digital photography exclusively or use both analog and digital media in their work. In either case, there is sure to be an archive of slides and negatives that cannot be directly integrated into the new digital workflow, nor can it be archived in a digital format. Increasingly, photographers are trying to bridge this gap with the use of high-performance film scanners. In this 2nd edition, you will learn how to achieve the best possible digital image from a negative or a slide, and how to build a workflow to make this process efficient, repeatable, and reliable. The author uses Nikon's film scanners, but all steps can easily be accomplished while using a different scanner. The most common software tools for scanning (SilverFast, VueScan, NikonScan) are not only covered extensively in the book, but trial versions are also provided on a DVD, which also contains other useful tools for image editing, as well as numerous sample scans.

  9. Classification of pulmonary emphysema from chest CT scans using integral geometry descriptors

    Science.gov (United States)

    van Rikxoort, E. M.; Goldin, J. G.; Galperin-Aizenberg, M.; Brown, M. S.

    2011-03-01

    To gain insight into the underlying pathways of emphysema and monitor the effect of treatment, methods to quantify and phenotype the different types of emphysema from chest CT scans are of crucial importance. Current standard measures rely on density thresholds for individual voxels, which is influenced by inspiration level and does not take into account the spatial relationship between voxels. Measures based on texture analysis do take the interrelation between voxels into account and therefore might be useful for distinguishing different types of emphysema. In this study, we propose to use Minkowski functionals combined with rotation invariant Gaussian features to distinguish between healthy and emphysematous tissue and classify three different types of emphysema. Minkowski functionals characterize binary images in terms of geometry and topology. In 3D, four Minkowski functionals are defined. By varying the threshold and size of neighborhood around a voxel, a set of Minkowski functionals can be defined for each voxel. Ten chest CT scans with 1810 annotated regions were used to train the method. A set of 108 features was calculated for each training sample from which 10 features were selected to be most informative. A linear discriminant classifier was trained to classify each voxel in the lungs into a subtype of emphysema or normal lung. The method was applied to an independent test set of 30 chest CT scans with varying amounts and types of emphysema with 4347 annotated regions of interest. The method is shown to perform well, with an overall accuracy of 95%.

  10. Comparison of dimensional accuracy of digital dental models produced from scanned impressions and scanned stone casts

    Science.gov (United States)

    Subeihi, Haitham

    Introduction: Digital models of dental arches play a more and more important role in dentistry. A digital dental model can be generated by directly scanning intraoral structures, by scanning a conventional impression of oral structures or by scanning a stone cast poured from the conventional impression. An accurate digital scan model is a fundamental part for the fabrication of dental restorations. Aims: 1. To compare the dimensional accuracy of digital dental models produced by scanning of impressions versus scanning of stone casts. 2. To compare the dimensional accuracy of digital dental models produced by scanning of impressions made of three different materials (polyvinyl siloxane, polyether or vinyl polyether silicone). Methods and Materials: This laboratory study included taking addition silicone, polyether and vinyl polyether silicone impressions from an epoxy reference model that was created from an original typodont. Teeth number 28 and 30 on the typodont with a missing tooth number 29 were prepared for a metal-ceramic three-unit fixed dental prosthesis with tooth #29 being a pontic. After tooth preparation, an epoxy resin reference model was fabricated by duplicating the typodont quadrant that included the tooth preparations. From this reference model 12 polyvinyl siloxane impressions, 12 polyether impressions and 12 vinyl polyether silicone impressions were made. All 36 impressions were scanned before pouring them with dental stone. The 36 dental stone casts were, in turn, scanned to produce digital models. A reference digital model was made by scanning the reference model. Six groups of digital models were produced. Three groups were made by scanning of the impressions obtained with the three different materials, the other three groups involved the scanning of the dental casts that resulted from pouring the impressions made with the three different materials. Groups of digital models were compared using Root Mean Squares (RMS) in terms of their

  11. Laryngotracheobronchial papillomatosis: findings on computed tomography scans of the chest

    International Nuclear Information System (INIS)

    Marchiori, Edson; Irion, Klaus Loureiro; Zanetti, Glaucia; Missrie, Israel; Sato, Juliana

    2008-01-01

    Objectives: To present the findings of computed tomography (CT) scans of the chest in patients with laryngotracheobronchial papillomatosis. Methods: We retrospectively analyzed CT scans of eight patients, five males and three females, ranging from 5 to 18 years of age with a mean age of 10.5 years. Images were independently reviewed by two radiologists. In discrepant cases, a consensus was reached. Results: The most common CT findings were intratracheal polypoid lesions and pulmonary nodules, many of which were cavitated. Conclusions: In patients with laryngotracheobronchial papillomatosis, the most common tomographic finding was the combination of intratracheal polypoid lesions and multiple pulmonary nodules, many of which were cavitated. (author)

  12. Laryngotracheobronchial papillomatosis: findings on computed tomography scans of the chest

    Energy Technology Data Exchange (ETDEWEB)

    Marchiori, Edson [Universidade Federal Fluminense (UFF), Niteroi, RJ (Brazil). Dept. de Radiologia; Araujo Neto, Cesar de [Universidade Federal da Bahia (UFBA), Salvador, BA (Brazil). Dept. de Radiologia; Meirelles, Gustavo Souza Portes [Universidade Federal de Sao Paulo (UNIFESP-EPM), SP (Brazil); Irion, Klaus Loureiro [The Royal Liverpool University Hospital, Liverpool (United Kingdom); Zanetti, Glaucia [Faculdade de Medicina de Petropolis (FMP), RJ (Brazil); Missrie, Israel [Universidade Federal de Sao Paulo (UNIFESP-EPM), SP (Brazil). Dept. de Diagnostico por Imagem; Sato, Juliana [Universidade Federal de Sao Paulo (UNIFESP-EPM), SP (Brazil). Dept. de Otorrinolaringologia e Cirurgia de Cabeca e Pescoco

    2008-12-15

    Objectives: To present the findings of computed tomography (CT) scans of the chest in patients with laryngotracheobronchial papillomatosis. Methods: We retrospectively analyzed CT scans of eight patients, five males and three females, ranging from 5 to 18 years of age with a mean age of 10.5 years. Images were independently reviewed by two radiologists. In discrepant cases, a consensus was reached. Results: The most common CT findings were intratracheal polypoid lesions and pulmonary nodules, many of which were cavitated. Conclusions: In patients with laryngotracheobronchial papillomatosis, the most common tomographic finding was the combination of intratracheal polypoid lesions and multiple pulmonary nodules, many of which were cavitated. (author)

  13. Interactive lung segmentation in abnormal human and animal chest CT scans

    International Nuclear Information System (INIS)

    Kockelkorn, Thessa T. J. P.; Viergever, Max A.; Schaefer-Prokop, Cornelia M.; Bozovic, Gracijela; Muñoz-Barrutia, Arrate; Rikxoort, Eva M. van; Brown, Matthew S.; Jong, Pim A. de; Ginneken, Bram van

    2014-01-01

    Purpose: Many medical image analysis systems require segmentation of the structures of interest as a first step. For scans with gross pathology, automatic segmentation methods may fail. The authors’ aim is to develop a versatile, fast, and reliable interactive system to segment anatomical structures. In this study, this system was used for segmenting lungs in challenging thoracic computed tomography (CT) scans. Methods: In volumetric thoracic CT scans, the chest is segmented and divided into 3D volumes of interest (VOIs), containing voxels with similar densities. These VOIs are automatically labeled as either lung tissue or nonlung tissue. The automatic labeling results can be corrected using an interactive or a supervised interactive approach. When using the supervised interactive system, the user is shown the classification results per slice, whereupon he/she can adjust incorrect labels. The system is retrained continuously, taking the corrections and approvals of the user into account. In this way, the system learns to make a better distinction between lung tissue and nonlung tissue. When using the interactive framework without supervised learning, the user corrects all incorrectly labeled VOIs manually. Both interactive segmentation tools were tested on 32 volumetric CT scans of pigs, mice and humans, containing pulmonary abnormalities. Results: On average, supervised interactive lung segmentation took under 9 min of user interaction. Algorithm computing time was 2 min on average, but can easily be reduced. On average, 2.0% of all VOIs in a scan had to be relabeled. Lung segmentation using the interactive segmentation method took on average 13 min and involved relabeling 3.0% of all VOIs on average. The resulting segmentations correspond well to manual delineations of eight axial slices per scan, with an average Dice similarity coefficient of 0.933. Conclusions: The authors have developed two fast and reliable methods for interactive lung segmentation in

  14. Digital luminescence radiography using a chest phantom

    International Nuclear Information System (INIS)

    Lyttkens, K.; Kehler, M.; Andersson, B.; Carlsen, S.; Ebbesen, A.; Hochbergs, P.; Stroembaeck, A.

    1993-01-01

    With the introduction of picture and archiving communicating systems an alternative image display for the wards might be a personal computer (PC). The intention with this study was to evaluate the diagnostic image quality of the monitor of a PC compared to that of a workstation. Eighty-five digital radiographs of a chest phantom with simulated tumors in the mediastinum and right lung were saved on optical discs. The examinations were reviewed by 4 radiologists on a monitor at a workstation and at a PC, and receiver operating characteristic (ROC) curves were constructed. No significant difference was found between performance of the PC and the workstation. (orig.)

  15. Comparison of chest radiography, chest digital tomosynthesis and low dose MDCT to detect small ground-glass opacity nodules: an anthropomorphic chest phantom study

    International Nuclear Information System (INIS)

    Doo, Kyung Won; Kang, Eun-Young; Yong, Hwan Seok; Ham, Soo-Youn; Lee, Ki Yeol; Choo, Ji Yung

    2014-01-01

    The purpose of this study was to evaluate the diagnostic performance of chest radiography (CXR), chest digital tomosynthesis (DT) and low dose multidetector computed tomography (LDCT) for the detection of small pulmonary ground-glass opacity (GGO) nodules, using an anthropomorphic chest phantom. Artificial pulmonary nodules were placed in a phantom and a total of 40 samples of different nodule settings underwent CXR, DT and LDCT. The images were randomly read by three experienced chest radiologists. Free-response receiver-operating characteristics (FROC) were used. The figures of merit for the FROC curves averaged for the three observers were 0.41, 0.37 and 0.76 for CXR, DT and LDCT, respectively. FROC analyses revealed significantly better performance of LDCT over CXR or DT for the detection of GGO nodules (P < 0.05). The difference in detectability between CXR and DT was not statistically significant (P = 0.73). The diagnostic performance of DT for the detection of pulmonary small GGO nodules was not significantly different from that of CXR, but LDCT performed significantly better than both CXR and DT. DT is not a suitable alternative to CT for small GGO nodule detection, and LDCT remains the method of choice for this purpose. (orig.)

  16. Comparison of chest radiography, chest digital tomosynthesis and low dose MDCT to detect small ground-glass opacity nodules: an anthropomorphic chest phantom study

    Energy Technology Data Exchange (ETDEWEB)

    Doo, Kyung Won; Kang, Eun-Young; Yong, Hwan Seok [Korea University Guro Hospital, Korea University College of Medicine, Department of Radiology, Seoul (Korea, Republic of); Ham, Soo-Youn [Korea University Anam Hospital, Korea University College of Medicine, Department of Radiology, Seoul (Korea, Republic of); Lee, Ki Yeol; Choo, Ji Yung [Korea University Ansan Hospital, Korea University College of Medicine, Department of Radiology, Ansan (Korea, Republic of)

    2014-12-15

    The purpose of this study was to evaluate the diagnostic performance of chest radiography (CXR), chest digital tomosynthesis (DT) and low dose multidetector computed tomography (LDCT) for the detection of small pulmonary ground-glass opacity (GGO) nodules, using an anthropomorphic chest phantom. Artificial pulmonary nodules were placed in a phantom and a total of 40 samples of different nodule settings underwent CXR, DT and LDCT. The images were randomly read by three experienced chest radiologists. Free-response receiver-operating characteristics (FROC) were used. The figures of merit for the FROC curves averaged for the three observers were 0.41, 0.37 and 0.76 for CXR, DT and LDCT, respectively. FROC analyses revealed significantly better performance of LDCT over CXR or DT for the detection of GGO nodules (P < 0.05). The difference in detectability between CXR and DT was not statistically significant (P = 0.73). The diagnostic performance of DT for the detection of pulmonary small GGO nodules was not significantly different from that of CXR, but LDCT performed significantly better than both CXR and DT. DT is not a suitable alternative to CT for small GGO nodule detection, and LDCT remains the method of choice for this purpose. (orig.)

  17. Evaluation of slot-to-slot coupling between dielectric slot waveguides and metal-insulator-metal slot waveguides.

    Science.gov (United States)

    Kong, Deqing; Tsubokawa, Makoto

    2015-07-27

    We numerically analyzed the power-coupling characteristics between a high-index-contrast dielectric slot waveguide and a metal-insulator-metal (MIM) plasmonic slot waveguide as functions of structural parameters. Couplings due mainly to the transfer of evanescent components in two waveguides generated high transmission efficiencies of 62% when the slot widths of the two waveguides were the same and 73% when the waveguides were optimized by slightly different widths. The maximum transmission efficiency in the slot-to-slot coupling was about 10% higher than that in the coupling between a normal slab waveguide and an MIM waveguide. Large alignment tolerance of the slot-to-slot coupling was also proved. Moreover, a small gap inserted into the interface between two waveguides effectively enhances the transmission efficiency, as in the case of couplings between a normal slab waveguide and an MIM waveguide. In addition, couplings with very wideband transmissions over a wavelength region of a few hundred nanometers were validated.

  18. Chest CT scans are frequently abnormal in asymptomatic patients with newly diagnosed acute myeloid leukemia.

    Science.gov (United States)

    Vallipuram, Janaki; Dhalla, Sidika; Bell, Chaim M; Dresser, Linda; Han, Heekyung; Husain, Shahid; Minden, Mark D; Paul, Narinder S; So, Miranda; Steinberg, Marilyn; Vallipuram, Mayuran; Wong, Gary; Morris, Andrew M

    2017-04-01

    Chest computed tomography (CT) findings of nodules, ground glass opacities, and consolidations are often interpreted as representing invasive fungal infection in individuals with febrile neutropenia. We assessed whether these CT findings were present in asymptomatic individuals with acute myeloid leukemia (AML) at low risk of invasive fungal disease. A retrospective study of consecutive asymptomatic adult patients with newly diagnosed AML over a 2-year period was performed at a tertiary care oncology center. Radiology reports of baseline chest CTs were reviewed. Of 145 CT scans, the majority (88%) had pulmonary abnormalities. Many (70%) had one or both of unspecified opacities (52%) and nodules (49%). Ground glass opacities (18%) and consolidations (12%) occurred less frequently. Radiologists suggested pneumonia as a possible diagnosis in 32% (n = 47) of scans. Chest CT may result in over-diagnosis of invasive fungal disease in individuals with febrile neutropenia if interpreted without correlation to the patients' clinical status.

  19. Chest tube placement in thorax trauma - comparison chest X-ray and computed tomography (CT)

    International Nuclear Information System (INIS)

    Heim, P.; Maas, R.; Buecheler, E.; Tesch, C.

    1998-01-01

    Estimation of chest tube placement in patients with thoracic trauma with regard to chest tube malposition in chest radiography in the supine position compared to additional computed tomography of the thorax. Material and methods: Apart from compulsory chest radiography after one or multiple chest tube insertions, 31 severely injured patients with thoracic trauma underwent a CT scan of the thorax. These 31 patients with 40 chest tubes constituted the basis for the present analysis. Results: In chest radiography in the supine position there were no chest tube malpositions (n=40); In the CT scans 25 correct positions, 7 pseudo-malpositions, 6 intrafissural and 2 intrapulmonary malpositions were identified. Moreover 16 sufficient, 18 insufficient and 6 indifferent functions of the chest tubes were seen. Conclusion: In case of lasting clinical problems and questionable function of the chest tube, chest radiography should be supplemented by a CT scan of the thorax in order to estimate the position of the chest tube. (orig.) [de

  20. Dose-image quality study in digital chest radiography using Monte Carlo simulation

    International Nuclear Information System (INIS)

    Correa, S.C.A.; Souza, E.M.; Silva, A.X.; Lopes, R.T.; Yoriyaz, H.

    2008-01-01

    One of the main preoccupations of diagnostic radiology is to guarantee a good image-sparing dose to the patient. In the present study, Monte Carlo simulations, with MCNPX code, coupled with an adult voxel female model (FAX) were performed to investigate how image quality and dose in digital chest radiography vary with tube voltage (80-150 kV) using air-gap technique and a computed radiography system. Calculated quantities were normalized to a fixed value of entrance skin exposure (ESE) of 0.0136 R. The results of the present analysis show that the image quality for chest radiography with imaging plate is improved and the dose reduced at lower tube voltage

  1. PC-based digital feedback control for scanning force microscope

    International Nuclear Information System (INIS)

    Mohd Ashhar Khalid

    2002-01-01

    In the past, most digital feedback implementation for scanned-probe microscope were based on a digital signal processor (DSP). At present DSP plug-in card with the input-output interface module is still expensive compared to a fast pentium PC motherboard. For a magnetic force microscope (MFM) digital feedback has an advantage where the magnetic signal can be easily separated from the topographic signal. In this paper, a simple low-cost PC-based digital feedback and imaging system for Scanning Force Microscope (SFM) is presented. (Author)

  2. Human Digital Modeling & Hand Scanning Lab

    Data.gov (United States)

    Federal Laboratory Consortium — This laboratory incorporates specialized scanning equipment, computer workstations and software applications for the acquisition and analysis of digitized models of...

  3. Effects of optimization and image processing in digital chest radiography

    International Nuclear Information System (INIS)

    Kheddache, S.; Maansson, L.G.; Angelhed, J.E.; Denbratt, L.; Gottfridsson, B.; Schlossman, D.

    1991-01-01

    A digital system for chest radiography based on a large image intensifier was compared to a conventional film-screen system. The digital system was optimized with regard to spatial and contrast resolution and dose. The images were digitally processed for contrast and edge enhancement. A simulated pneumothorax and two and two simulated nodules were positioned over the lungs and the mediastinum of an anthro-pomorphic phantom. Observer performance was evaluated with Receiver Operating Characteristic (ROC) analysis. Five observers assessed the processed digital images and the conventional full-size radiographs. The time spent viewing the full-size radiographs and the digital images was recorded. For the simulated pneumothorax, the results showed perfect performance for the full-size radiographs and detectability was high also for the processed digital images. No significant differences in the detectability of the simulated nodules was seen between the two imaging systems. The results for the digital images showed a significantly improved detectability for the nodules in the mediastinum as compared to a previous ROC study where no optimization and image processing was available. No significant difference in detectability was seen between the former and the present ROC study for small nodules in the lung. No difference was seen in the time spent assessing the conventional full-size radiographs and the digital images. The study indicates that processed digital images produced by a large image intensifier are equal in image quality to conventional full-size radiographs for low-contrast objects such as nodules. (author). 38 refs.; 4 figs.; 1 tab

  4. Birt-Hogg-Dube syndrome prospectively detected by review of chest computed tomography scans.

    Directory of Open Access Journals (Sweden)

    Hye Jung Park

    Full Text Available Birt-Hogg-Dube syndrome (BHD is a rare disorder caused by mutations in the gene that encodes folliculin (FLCN and is inherited in an autosomal dominant manner. BHD is commonly accompanied by fibrofolliculomas, renal tumors, multiple pulmonary cysts, and spontaneous pneumothorax. The aim of this study was to detect BHD prospectively in patients undergoing chest computed tomography (CT scans and to evaluate further the characteristics of BHD in Korea.We prospectively checked and reviewed the chest CT scans obtained for 10,883 patients at Gangnam Severance Hospital, Seoul, Korea, from June 1, 2015 to May 31, 2016. Seventeen patients met the study inclusion criteria and underwent screening for FLCN mutation to confirm BHD. We analyzed the characteristics of the patients confirmed to have BHD and those for a further 6 patients who had previously been described in Korea.Six (0.06% of the 10,883 patients reviewed were diagnosed with BHD. There was no difference in demographic or clinical features between the patients with BHD (n = 6 and those without BHD (n = 11. Pneumothorax was present in 50% of the patients with BHD but typical skin and renal lesions were absent. The maximum size of the cysts in the BHD group (median 39.4 mm; interquartile range [IQR] 11.4 mm was significantly larger than that in the non-BHD group (median 15.8 mm; IQR 7.8 mm; P = 0.001. Variable morphology was seen in 100.0% of the cysts in the BHD group but in only 18.2% of the cysts in the non-BHD group (P = 0.002. Nine (95% of the total of 12 Korean patients with BHD had experienced pneumothorax. Typical skin and renal lesions were present in 20.0% of patients with BHD.Our findings suggest that BHD can be detected if chest CT scans are read in detail.

  5. The measurement of organic radiation dose of multi-slice CT scanning by using the Chinese anthropomorphic chest phantom

    International Nuclear Information System (INIS)

    Peng Gang; Zeng Yongming; Luo Tianyou; Zhao Feng; Zhang Zhiwei; Yu Renqiang; Peng Shengkun

    2011-01-01

    Objective: Using the Chinese anthropomorphic chest phantom to measure the absorbed dose of various tissues and organs under different noise index, and to assess the radiation dose of MSCT chest scanning with the effective dose (ED). Methods: The equivalence of the Chinese anthropomorphic chest phantom (CDP-1 C) and the adult chest on CT sectional anatomy and X-ray attenuation was demonstrated. The absorbed doses of various tissues and organs under different noise index were measured by laying thermoluminescent dosimeters (TLD) inside the phantom, and the corresponding dose-length products (DLP) were recorded. Both of them were later converted into ED and comparison was conducted to analyze the dose levels of chest CT scanning with automatic tube current modulation (ATCM) under different noise index. Student t-test was applied using SPSS 12.0 statistical software. Results: The Phantom was similar to the human body on CT sectional anatomy. The average CT value of phantom are - 788.04 HU in lung, 45.64 HU in heart, 65.84 HU in liver, 254.32 HU in spine and the deviations are 0.10%, 3.04%, 4.49% and 4.36% respectively compared to humans. The difference of average CT value of liver was statistically significant (t=-8.705, P 0.05). As the noise index increased from 8.5 to 22.5, the DLP decreased from 393.57 mGy · cm to 78.75 mGy · cm and the organs dose declined. For example, the average absorbed dose decreased from 22.38 mGy to 3.66 mGy in lung. Compared to ED calculating by absorbed dose, the ED calculating by DLP was lower. The ED values of the two methods were 6.69 mSv and 8.77 mSv when the noise index was set at 8.5. Conclusions: Application of the Chinese anthropomorphic chest phantom to carry out CT dose assessment is more accurate. The noise index should be set more than 8.5 during the chest CT scanning based on ATCM technique. (authors)

  6. Out-of-Print Digital Scanning.

    Science.gov (United States)

    Kellerman, L. Suzanne

    2002-01-01

    Describes the in-house operational procedures developed at the Penn State University libraries to produce facsimiles of hard-to-locate, out-of-print titles using digital scanning technologies. Highlights include identification and selection, costs, copyright issues, interlibrary loan, materials preparation, and future trends. (Author/LRW)

  7. Dosimetric study in chest computed tomography scans of adult and pediatric phantoms

    International Nuclear Information System (INIS)

    Namen A, W.; Prata M, A.; Guedes, G.

    2016-10-01

    The computed tomography scan is a radiological technique that permits an evaluation of the patient internal structures. In the last ten years, this technique has had a high growth due to clinical cases of medical emergencies, cancer and pediatric trauma. Widespread of this technique has a significant increase in the patient dose. The risk associated with the radiological examination can be considered very low compared to the natural risk. However, any additional risk, no matter how small, is unacceptable if it does not benefit the patient. To be aware of the dose distribution is important when the objective is to vary the acquisition parameters aiming a dose reduction. The aim os this study is develop a pediatric chest phantom to evaluate the dose variation in CT scans. In this work, a cylindrical adult chest phantom made in polymethyl methacrylate was used and a second chest phantom was developed, based on dimensions of in eight year old patient in oblong shape. The two simulators have 5 openings, one is central and four are peripheral lagged by 90 degrees Celsius, which allow positioning a pencil chamber aiming and observation of the dose in 5 regions. In a GE CT scanner, Discovery model and 64 channels, the central slice of both simulators were irradiated successively to obtain dose measurements using a pencil chamber. The irradiation of the central slice was conducted using the service protocol. The registered dose values showed that the pediatric phantom had higher doses especially in the anterior, posterior and central regions. The results also enabled a comparison among the index dose values obtained from the measurements with the pencil chamber. (Author)

  8. Dosimetric study in chest computed tomography scans of adult and pediatric phantoms

    Energy Technology Data Exchange (ETDEWEB)

    Namen A, W.; Prata M, A. [Universidade Federal de Minas Gerais, Departamento de Engenharia Nuclear, Av. Pte. Antonio Carlos 6627, 31270-901 Pampulha, Belo Horizonte, Minas Gerais (Brazil); Guedes, G., E-mail: wadia.namen@gmail.com [Centro Federal de Educacao Tecnologica de Minas Gerais, Centro de Engenharia Biomedica, Av. Amazonas 5253, 30421-169 Nova Suica, Belo Horizonte, Minas Gerais (Brazil)

    2016-10-15

    The computed tomography scan is a radiological technique that permits an evaluation of the patient internal structures. In the last ten years, this technique has had a high growth due to clinical cases of medical emergencies, cancer and pediatric trauma. Widespread of this technique has a significant increase in the patient dose. The risk associated with the radiological examination can be considered very low compared to the natural risk. However, any additional risk, no matter how small, is unacceptable if it does not benefit the patient. To be aware of the dose distribution is important when the objective is to vary the acquisition parameters aiming a dose reduction. The aim os this study is develop a pediatric chest phantom to evaluate the dose variation in CT scans. In this work, a cylindrical adult chest phantom made in polymethyl methacrylate was used and a second chest phantom was developed, based on dimensions of in eight year old patient in oblong shape. The two simulators have 5 openings, one is central and four are peripheral lagged by 90 degrees Celsius, which allow positioning a pencil chamber aiming and observation of the dose in 5 regions. In a GE CT scanner, Discovery model and 64 channels, the central slice of both simulators were irradiated successively to obtain dose measurements using a pencil chamber. The irradiation of the central slice was conducted using the service protocol. The registered dose values showed that the pediatric phantom had higher doses especially in the anterior, posterior and central regions. The results also enabled a comparison among the index dose values obtained from the measurements with the pencil chamber. (Author)

  9. Automatic method for selective enhancement of different tissue densities at digital chest radiography

    International Nuclear Information System (INIS)

    McNitt-Gray, M.F.; Taira, R.K.; Eldredge, S.L.; Razavi, M.

    1991-01-01

    This paper reports that digital chest radiographs often are too bright and/or lack contrast when viewed on a video display. The authors have developed a method that can automatically provide a series of look-up tables that selectively enhance the radiographically soft or dense tissues on a digital chest radiograph. This reduces viewer interaction and improves displayed image quality. On the basis of a histogram analysis, gray-level ranges are approximated for the patient background, radiographically soft tissues, and radiographically dense tissues. A series of look-up tables is automatically created by varying the contrast in each range to achieve a level of enhancement for a selected tissue range. This is repeated for differing amounts of enhancement and for each tissue range. This allows the viewer to interactively select a tissue density range and degree of enhancement at the time of display via precalculated look-up tables. Preclinical trials in pediatric radiology using computed radiography images show that this method reduces viewer interaction and improves or maintains the displayed image quality

  10. Integration of Digital Dental Casts in Cone-Beam Computed Tomography Scans

    Science.gov (United States)

    Rangel, Frits A.; Maal, Thomas J. J.; Bergé, Stefaan J.; Kuijpers-Jagtman, Anne Marie

    2012-01-01

    Cone-beam computed tomography (CBCT) is widely used in maxillofacial surgery. The CBCT image of the dental arches, however, is of insufficient quality to use in digital planning of orthognathic surgery. Several authors have described methods to integrate digital dental casts into CBCT scans, but all reported methods have drawbacks. The aim of this feasibility study is to present a new simplified method to integrate digital dental casts into CBCT scans. In a patient scheduled for orthognathic surgery, titanium markers were glued to the gingiva. Next, a CBCT scan and dental impressions were made. During the impression-taking procedure, the titanium markers were transferred to the impression. The impressions were scanned, and all CBCT datasets were exported in DICOM format. The two datasets were matched, and the dentition derived from the scanned impressions was transferred to the CBCT of the patient. After matching the two datasets, the average distance between the corresponding markers was 0.1 mm. This novel method allows for the integration of digital dental casts into CBCT scans, overcoming problems such as unwanted extra radiation exposure, distortion of soft tissues due to the use of bite jigs, and time-consuming digital data handling. PMID:23050159

  11. Evaluation of variation of voltage (kV) absorbed dose in chest CT scans

    International Nuclear Information System (INIS)

    Mendonca, Bruna G.A.; Mourao, Arnaldo P.

    2013-01-01

    Computed tomography (CT) is one of the most important diagnostic techniques images today. The increasing utilization of CT implies a significant increase of population exposure to ionizing radiation. Optimization of practice aims to reduce doses to patients because the image quality is directly related to the diagnosis. You can decrease the amount of dose to the patient, and maintain the quality of the image. There are several parameters that can be manipulated in a CT scan and these parameters can be used to reduce the energy deposited in the patient. Based on this, we analyzed the variation of dose deposited in the lungs, breasts and thyroid, by varying the supply voltage of the tube. Scans of the thorax were performed following the protocol of routine chest with constant and variable current for the same applied voltage. Moreover, a female phantom was used and thermoluminescent dosimeters (TLD-100), model bat, were used to record the specific organ doses. Scans were performed on a GE CT scanner, model 64 Discovery channels. Higher doses were recorded for the voltage of 120 kV with 200 mAs in the lungs (22.46 mGy) and thyroid (32.22 mGy). For scans with automatic mAs, variable between 100 and 440, this same tension contributed to the higher doses. The best examination in terms of the dose that was used with automatic 80 kV mAs, whose lungs and thyroid received lower dose. For the best breast exam was 100 kV. Since the increase in the 80 kV to 100 kV no impact so much the dose deposited in the lungs, it can be concluded that lowering the applied voltage to 100 kV resulted in a reduction in the dose absorbed by the patient. These results can contribute to optimizing scans of the chest computed tomography

  12. Radiological diagnosis of chest wall tuberculosis: CT versus chest radiograph

    International Nuclear Information System (INIS)

    Liu Fugeng; Pan Jishu; Chen Qihang; Zhou Cheng; Yu Jingying; Tang Dairong

    2006-01-01

    Objective: To evaluate the role of CT or Chest radiograph in diagnosis of chest wall tuberculosis. Methods: The study population included 21 patients with chest wall tuberculosis confirmed by operation or biopsy. Chest radiograph and plain CT were performed in all eases, while enhanced CT in 9 cases, and all images were reviewed by 2 radiologists. Results: Single soft tissue mass of the chest wall was detected in all cases on CT, but not on chest radiograph(χ 2 =42.000, P 2 =4.421, P<0.05). Conclusion: CT, especially enhanced CT scan is the first choice in the diagnosis of chest wall tuberculosis. (authors)

  13. Accuracy of chest radiography versus chest computed tomography in hemodynamically stable patients with blunt chest trauma

    Directory of Open Access Journals (Sweden)

    Chardoli Mojtaba

    2013-12-01

    Full Text Available 【Abstract】 Objective: Thoracic injuries are respon- sible for 25% of deaths of blunt traumas. Chest X-ray (CXR is the first diagnostic method in patients with blunt trauma. The aim of this study was to detect the accuracy of CXR versus chest computed tomograpgy (CT in hemodynami- cally stable patients with blunt chest trauma. Methods: Study was conducted at the emergency department of Sina Hospital from March 2011 to March 2012. Hemodynamically stable patients with at least 16 years of age who had blunt chest trauma were included. All patients underwent the same diagnostic protocol which consisted of physical examination, CXR and CT scan respectively. Results: Two hundreds patients (84% male and 16% female were included with a mean age of (37.9±13.7 years. Chin J Traumatol 2013;16(6:351-354 Rib fracture was the most common finding of CXR (12.5% and CT scan (25.5%. The sensitivity of CXR for hemothorax, thoracolumbar vertebra fractures and rib fractures were 20%, 49% and 49%, respectively. Pneumothorax, foreign body, emphysema, pulmonary contusion, liver hematoma and ster- num fracture were not diagnosed with CXR alone. Conclusion: Applying CT scan as the first-line diag- nostic modality in hemodynamically stable patients with blunt chest trauma can detect pathologies which may change management and outcome. Key words: Radiography; Thoracic injuries; Tomography, X-ray computed

  14. Anatomic and energy variation of scatter compensation for digital chest radiography with Fourier deconvolution

    International Nuclear Information System (INIS)

    Floyd, C.E.; Beatty, P.T.; Ravin, C.E.

    1988-01-01

    The Fourier deconvolution algorithm for scatter compensation in digital chest radiography has been evaluated in four anatomically different regions at three energies. A shift invariant scatter distribution shape, optimized for the lung region at 140 kVp, was applied at 90 kVp and 120 kVp in the lung, retrocardiac, subdiaphragmatic, and thoracic spine regions. Scatter estimates from the deconvolution were compared with measured values. While some regional variation is apparent, the use of a shift invariant scatter distribution shape (optimized for a given energy) produces reasonable scatter compensation in the chest. A different set of deconvolution parameters were required at the different energies

  15. Pediatric digital chest imaging.

    Science.gov (United States)

    Tarver, R D; Cohen, M; Broderick, N J; Conces, D J

    1990-01-01

    The Philips Computed Radiography system performs well with pediatric portable chest radiographs, handling the throughout of a busy intensive care service 24 hours a day. Images are excellent and routinely provide a conventional (unenhanced) image and an edge-enhanced image. Radiation dose is decreased by the lowered frequency of repeat examinations and the ability of the plates to respond to a much lower dose and still provide an adequate image. The high quality and uniform density of serial PCR portable radiographs greatly enhances diagnostic content of the films. Decreased resolution has not been a problem clinically. Image manipulation and electronic transfer to remote viewing stations appear to be helpful and are currently being evaluated further. The PCR system provides a marked improvement in pediatric portable chest radiology.

  16. Pediatric digital chest imaging

    International Nuclear Information System (INIS)

    Tarver, R.D.; Cohen, M.; Broderick, N.J.; Conces, D.J. Jr.

    1990-01-01

    The Philips Computed Radiography system performs well with pediatric portable chest radiographs, handling the throughout of a busy intensive care service 24 hours a day. Images are excellent and routinely provide a conventional (unenhanced) image and an edge-enhanced image. Radiation dose is decreased by the lowered frequency of repeat examinations and the ability of the plates to respond to a much lower dose and still provide an adequate image. The high quality and uniform density of serial PCR portable radiographs greatly enhances diagnostic content of the films. Decreased resolution has not been a problem clinically. Image manipulation and electronic transfer to remote viewing stations appear to be helpful and are currently being evaluated further. The PCR system provides a marked improvement in pediatric portable chest radiology

  17. Preliminary clinical evaluation of hard- and soft-copy digitized chest radiography

    Science.gov (United States)

    Rian, Roger L.; Smerud, Michael J.; Guinn, Todd

    1994-05-01

    The digital applications in radiology are a controversial advanced which potentially will influence all areas of patient imaging. It is utilized and accepted in angiography, computed tomography, magnetic resonance, nuclear imaging and sonography. More recently Computed Radiography has gained credibility in mobile scenarios as well as specific applications from cervical spine radiography to digital fluoroscopy. Usually this acceptance is related to benefits of lesser radiation exposure or an improved presentation with an incorrect radiographic technique. One advantage of interpreting from digital information is the potential manipulation of the image presentation to the observer through windowing, leveling and edge enhancement pre and/or during image review. Additionally this digital data can be transmitted over distance and represented as hard and/or soft copy for primary or consultative review. The number and quality of the images to be viewed, the environment of the review station as well as the observer experience with conventional radiographic as well as digital image evaluation are important aspects of delivering the radiologist's product i.e. the final interpretation. This paper assesses that product, specifically addressing the question `Is the radiologist's report the same whether derived from the original analog image or from its digitized image.' The object of this study is to determine whether a digital system (3M PACS) designed for consultative viewing in a satellite department can also be used directly for primary diagnosis of conventional chest exams.

  18. Clinical value of CARE dose 4D technique in decreasing CT scanning dose of adult chest

    International Nuclear Information System (INIS)

    Wu Aiqin; Zheng Wenlong; Xu Chongyong; Fang Bidong; Ge Wen

    2011-01-01

    Objective: To investigate the value of CARE Dose 4D technique in decreasing radiation dose and improving image quality of multi-slice spiral CT in adult chest scanning. Methods: 100 patients of chest CT scanning were equally divided into study group and control group randomly. CARE Dose 4D Technique was used in study group. Effective mAs value, volume CT dose index (CTDI vol ) and dose length product (DLP) were displayed automatically in machine while chest scanning; those values and actual mAs value of every image were recorded respectively. The image quality at apex of lung, lower edge of aorta arch, middle area of left atrium and base of lung on every image of 400 images was judged and classified as three level (excellent, good, poor) by two deputy chief physicians with double blind method, the image noise at corresponding parts was measured. Results: While setting 80 mAs for quality reference mAs, the effective mAs value in study group most decreased 44 mAs than control group with an average decrease of 9.60 (12.0%), CTDI vol with 4.75 mGy with an average decrease of 0.95 mCy (11.0%), DLP 99.50% in study group, with 98.0% in control group. But it was higher at apex of lung and base of lung, lower at middle area of left atrium, and similar at lower edge of aorta arch in study group than contrast group. The image noise were lower at apex of lung and base of lung in study group than control group (t =6.299 and 2.332, all P<0.05), higher at middle area of left atrium in study group than control group (t=3.078, P<0.05) and similar at lower edge of aorta arch in study group than control group (t=1.191, P>0.05). Conclusions: CARE Dose 4D technique provides a function regulated mAs real-time on line, it not only raises utilization rate of radiation and decreases radiation dose, but also promises and increases image quality in chest CT scanning, and has some clinical significance. (authors)

  19. Anteroposterior chest radiograph vs. chest CT scan in early detection of pneumothorax in trauma patients

    OpenAIRE

    Omar, Hesham R; Mangar, Devanand; Khetarpal, Suneel; Shapiro, David H; Kolla, Jaya; Rashad, Rania; Helal, Engy; Camporesi, Enrico M

    2011-01-01

    Abstract Pneumothorax is a common complication following blunt chest wall trauma. In these patients, because of the restrictions regarding immobilization of the cervical spine, Anteroposterior (AP) chest radiograph is usually the most feasible initial study which is not as sensitive as the erect chest X-ray or CT chest for detection of a pneumothorax. We will present 3 case reports which serve for better understanding of the entity of occult pneumothorax. The first case is an example of a tru...

  20. Chest CT in children: anesthesia and atelectasis

    Energy Technology Data Exchange (ETDEWEB)

    Newman, Beverley; Gawande, Rakhee [Lucile Packard Children' s Hospital, Department of Radiology, Stanford, CA (United States); Krane, Elliot J. [Stanford University School of Medicine, Lucile Packard Children' s Hospital, Department of Anesthesiology, Perioperative and Pain Medicine, Stanford, CA (United States); Holmes, Tyson H. [Stanford University School of Medicine, Lucile Packard Children' s Hospital, Department of Psychiatry and Behavioral Sciences, Stanford, CA (United States); Robinson, Terry E. [Stanford University School of Medicine, Lucile Packard Children' s Hospital, Department of Pulmonary Medicine and Cystic Fibrosis Center for Excellence in Pulmonary Biology, Stanford, CA (United States)

    2014-02-15

    There has been an increasing tendency for anesthesiologists to be responsible for providing sedation or anesthesia during chest CT imaging in young children. Anesthesia-related atelectasis noted on chest CT imaging has proven to be a common and troublesome problem, affecting image quality and diagnostic sensitivity. To evaluate the safety and effectiveness of a standardized anesthesia, lung recruitment, controlled-ventilation technique developed at our institution to prevent atelectasis for chest CT imaging in young children. Fifty-six chest CT scans were obtained in 42 children using a research-based intubation, lung recruitment and controlled-ventilation CT scanning protocol. These studies were compared with 70 non-protocolized chest CT scans under anesthesia taken from 18 of the same children, who were tested at different times, without the specific lung recruitment and controlled-ventilation technique. Two radiology readers scored all inspiratory chest CT scans for overall CT quality and atelectasis. Detailed cardiorespiratory parameters were evaluated at baseline, and during recruitment and inspiratory imaging on 21 controlled-ventilation cases and 8 control cases. Significant differences were noted between groups for both quality and atelectasis scores with optimal scoring demonstrated in the controlled-ventilation cases where 70% were rated very good to excellent quality scans compared with only 24% of non-protocol cases. There was no or minimal atelectasis in 48% of the controlled ventilation cases compared to 51% of non-protocol cases with segmental, multisegmental or lobar atelectasis present. No significant difference in cardiorespiratory parameters was found between controlled ventilation and other chest CT cases and no procedure-related adverse events occurred. Controlled-ventilation infant CT scanning under general anesthesia, utilizing intubation and recruitment maneuvers followed by chest CT scans, appears to be a safe and effective method to obtain

  1. Chest CT in children: anesthesia and atelectasis

    International Nuclear Information System (INIS)

    Newman, Beverley; Gawande, Rakhee; Krane, Elliot J.; Holmes, Tyson H.; Robinson, Terry E.

    2014-01-01

    There has been an increasing tendency for anesthesiologists to be responsible for providing sedation or anesthesia during chest CT imaging in young children. Anesthesia-related atelectasis noted on chest CT imaging has proven to be a common and troublesome problem, affecting image quality and diagnostic sensitivity. To evaluate the safety and effectiveness of a standardized anesthesia, lung recruitment, controlled-ventilation technique developed at our institution to prevent atelectasis for chest CT imaging in young children. Fifty-six chest CT scans were obtained in 42 children using a research-based intubation, lung recruitment and controlled-ventilation CT scanning protocol. These studies were compared with 70 non-protocolized chest CT scans under anesthesia taken from 18 of the same children, who were tested at different times, without the specific lung recruitment and controlled-ventilation technique. Two radiology readers scored all inspiratory chest CT scans for overall CT quality and atelectasis. Detailed cardiorespiratory parameters were evaluated at baseline, and during recruitment and inspiratory imaging on 21 controlled-ventilation cases and 8 control cases. Significant differences were noted between groups for both quality and atelectasis scores with optimal scoring demonstrated in the controlled-ventilation cases where 70% were rated very good to excellent quality scans compared with only 24% of non-protocol cases. There was no or minimal atelectasis in 48% of the controlled ventilation cases compared to 51% of non-protocol cases with segmental, multisegmental or lobar atelectasis present. No significant difference in cardiorespiratory parameters was found between controlled ventilation and other chest CT cases and no procedure-related adverse events occurred. Controlled-ventilation infant CT scanning under general anesthesia, utilizing intubation and recruitment maneuvers followed by chest CT scans, appears to be a safe and effective method to obtain

  2. Regionally adaptive histogram equalization of the chest

    International Nuclear Information System (INIS)

    Sherrier, R.H.; Johnson, G.A.

    1986-01-01

    Advances in digital chest radiography have resulted in the acquisition of high-quality digital images of the human chest. With these advances, there arises a genuine need for image processing algorithms, specific to chest images. The author has implemented the technique of histogram equalization, noting the problems encountered when it is adapted to chest images. These problems have been successfully solved with a regionally adaptive histogram equalization method. Histograms are calculated locally and then modified according to both the mean pixel value of a given region and certain characteristics of the cumulative distribution function. The method allows certain regions of the chest radiograph to be enhanced differentially

  3. A comparison of digital tomosynthesis and chest radiography in evaluating airway lesions using computed tomography as a reference.

    Science.gov (United States)

    Choo, Ji Yung; Lee, Ki Yeol; Yu, Ami; Kim, Je-Hyeong; Lee, Seung Heon; Choi, Jung Won; Kang, Eun-Young; Oh, Yu Whan

    2016-09-01

    To compare the diagnostic performance of digital tomosynthesis (DTS) and chest radiography for detecting airway abnormalities, using computed tomography (CT) as a reference. We evaluated 161 data sets from 149 patients (91 with and 70 without airway abnormalities) who had undergone radiography, DTS, and CT to detect airway problems. Radiographs and DTS were evaluated to localize and score the severity of the airway abnormalities, and to score the image quality using CT as a reference. Receiver operating characteristics (ROC), McNemar's test, weighted kappa, and the paired t-test were used for statistical analysis. The sensitivity of DTS was higher (reader 1, 93.51 %; reader 2, 94.29 %) than chest radiography (68.83 %; 71.43 %) in detecting airway lesions. The diagnostic accuracy of DTS (90.91 %; 94.70 %) was also significantly better than that of radiography (78.03 %; 82.58 %, all p chest radiography (1.83, 2.74; p < 0.05) in the results of both readers. The inter-observer agreement with respect to DTS findings was moderate and superior when compared to radiography findings. DTS is a more accurate and sensitive modality than radiography for detecting airway lesions that are easily obscured by soft tissue structures in the mediastinum. • Digital tomosynthesis offers new diagnostic options for airway lesions. • Digital tomosynthesis is more sensitive and accurate than radiography for airway lesions. • Digital tomosynthesis shows better image quality than radiography. • Assessment of lesion severity, via tomosynthesis is comparable to computed tomography.

  4. Integration of Digital Dental Casts in Cone-Beam Computed Tomography Scans

    OpenAIRE

    Rangel, Frits A.; Maal, Thomas J. J.; Bergé, Stefaan J.; Kuijpers-Jagtman, Anne Marie

    2012-01-01

    Cone-beam computed tomography (CBCT) is widely used in maxillofacial surgery. The CBCT image of the dental arches, however, is of insufficient quality to use in digital planning of orthognathic surgery. Several authors have described methods to integrate digital dental casts into CBCT scans, but all reported methods have drawbacks. The aim of this feasibility study is to present a new simplified method to integrate digital dental casts into CBCT scans. In a patient scheduled for orthognathic ...

  5. Using 99mTc-DTPA radioaerosol inhalation lung scan as compared with computed tomography to detect lung injury in blunt chest trauma

    International Nuclear Information System (INIS)

    Esme, H.; Kaya, E.; Solak, O.; Yavuz, Y.; Yurumez, Y.; Sezer, M.

    2007-01-01

    Detection of pulmonary contusion in patients with blunt chest trauma is very important so as to commence therapy immediately to avoid irreversible damage. The purpose of our study was to evaluate the efficacy of technetium-99m diethylene triamine penta-acetic acid ( 99m Tc-DTPA) aerosol inhalation lung scintigraphy in comparison with chest computed tomography (CT) in the diagnosis of pulmonary contusion at acute blunt chest trauma. Twenty-nine patients with isolated blunt chest trauma were referred to the emergency department of our hospital, and nine healthy people participated in this study. Sixteen patients who had pulmonary contusion on CT scans were referred to as group 1, and 13 patients who had normal CT scans as group 2. Nine healthy people comprised a control group. 99m Tc-DTPA aerosol inhalation lung scintigraphy was performed on the first day in all patients. The mean half time (T 1/2 ) and penetration index values of 99m Tc-DTPA clearance were significantly lower in groups 1 and 2 compared with the control group. Among the three groups, there were no significant differences in arterial blood gas analysis except for PO 2 . The mean T 1/2 value of 99m Tc-DTPA clearance did correlate with PO 2 values but not with pH, PCO 2 , or HCO 3 values. 99m Tc-DTPA radioaerosol inhalation lung imaging may serve as a useful adjunct and supportive method to chest CT scanning for detecting mild pulmonary contusion. (author)

  6. Surpassing digital holography limits by lensless object scanning holography.

    Science.gov (United States)

    Micó, Vicente; Ferreira, Carlos; García, Javier

    2012-04-23

    We present lensless object scanning holography (LOSH) as a fully lensless method, capable of improving image quality in reflective digital Fourier holography, by means of an extremely simplified experimental setup. LOSH is based on the recording and digital post-processing of a set of digital lensless holograms and results in a synthetic image with improved resolution, field of view (FOV), signal-to-noise ratio (SNR), and depth of field (DOF). The superresolution (SR) effect arises from the generation of a synthetic aperture (SA) based on the linear movement of the inspected object. The same scanning principle enlarges the object FOV. SNR enhancement is achieved by speckle suppression and coherent artifacts averaging due to the coherent addition of the multiple partially overlapping bandpass images. And DOF extension is performed by digital refocusing to different object's sections. Experimental results showing an impressive image quality improvement are reported for a one-dimensional reflective resolution test target. © 2012 Optical Society of America

  7. Transport Infrastructure Slot Allocation

    NARCIS (Netherlands)

    Koolstra, K.

    2005-01-01

    In this thesis, transport infrastructure slot allocation has been studied, focusing on selection slot allocation, i.e. on longer-term slot allocation decisions determining the traffic patterns served by infrastructure bottlenecks, rather than timetable-related slot allocation problems. The

  8. Does the usage of digital chest drainage systems reduce pleural inflammation and volume of pleural effusion following oncologic pulmonary resection?-A prospective randomized trial.

    Science.gov (United States)

    De Waele, Michèle; Agzarian, John; Hanna, Waël C; Schieman, Colin; Finley, Christian J; Macri, Joseph; Schneider, Laura; Schnurr, Terri; Farrokhyar, Forough; Radford, Katherine; Nair, Parameswaran; Shargall, Yaron

    2017-06-01

    Prolonged air leak and high-volume pleural drainage are the most common causes for delays in chest tube removal following lung resection. While digital pleural drainage systems have been successfully used in the management of post-operative air leak, their effect on pleural drainage and inflammation has not been studied before. We hypothesized that digital drainage systems (as compared to traditional analog continuous suction), using intermittent balanced suction, are associated with decreased pleural inflammation and postoperative drainage volumes, thus leading to earlier chest tube removal. One hundred and three [103] patients were enrolled and randomized to either analog (n=50) or digital (n=53) drainage systems following oncologic lung resection. Chest tubes were removed according to standardized, pre-defined protocol. Inflammatory mediators [interleukin-1B (IL-1B), 6, 8, tumour necrosis factor-alpha (TNF-α)] in pleural fluid and serum were measured and analysed. The primary outcome of interest was the difference in total volume of postoperative fluid drainage. Secondary outcome measures included duration of chest tube in-situ, prolonged air-leak incidence, length of hospital stay and the correlation between pleural effusion formation, degree of inflammation and type of drainage system used. There was no significant difference in total amount of fluid drained or length of hospital stay between the two groups. A trend for shorter chest tube duration was found with the digital system when compared to the analog (P=0.055). Comparison of inflammatory mediator levels revealed no significant differences between digital and analog drainage systems. The incidence of prolonged post-operative air leak was significantly higher when using the analog system (9 versus 2 patients; P=0.025). Lobectomy was associated with longer chest tube duration (P=0.001) and increased fluid drainage when compared to sub-lobar resection (Pdigital drainage does not appear to decrease pleural

  9. Automatic anatomically selective image enhancement in digital chest radiography

    International Nuclear Information System (INIS)

    Sezan, M.I.; Minerbo, G.N.; Schaetzing, R.

    1989-01-01

    The authors develop a technique for automatic anatomically selective enhancement of digital chest radiographs. Anatomically selective enhancement is motivated by the desire to simultaneously meet the different enhancement requirements of the lung field and the mediastinum. A recent peak detection algorithm and a set of rules are applied to the image histogram to determine automatically a gray-level threshold between the lung field and mediastinum. The gray-level threshold facilitates anatomically selective gray-scale modification and/or unsharp masking. Further, in an attempt to suppress possible white-band or black-band artifacts due to unsharp masking at sharp edges, local-contrast adaptivity is incorporated into anatomically selective unsharp masking by designing an anatomy-sensitive emphasis parameter which varies asymmetrically with positive and negative values of the local image contrast

  10. Automatic, anatomically selective, artifact-free enhancement of digital chest radiographs

    International Nuclear Information System (INIS)

    Sezan, M.I.; Tekalp, A.M.; Schaetzing, R.

    1988-01-01

    The authors propose a technique for automatic, anatomically selective, artifact-free enhancement of digital chest radiographs. Anatomically selective enhancement is motivated by the different enhancement requirements of the lung field and the mediastinum. A recent peak detection algorithm is applied to the image histogram to automatically determine a gray-level threshold between the lung and mediastinum fields. The gray-level threshold facilitates anatomically selective gray-scale modification and unsharp masking. Further, in an attempt to suppress possible white-band artifacts due to unsharp masking at sharp edges, local-contrast adaptivity is incorporated into anatomically selective unsharp masking by designing an anatomy-sensitive emphasis parameter that varied asymmetrically with positive and negative values of the local image contrast

  11. An Evaluation of pectus excavatum by means of body surface measurements on chest CT scans

    International Nuclear Information System (INIS)

    Hirotani, Taichi; Ohama, Kazunori; Shimotake, Takashi; Ishikawa, Nobuki; Watanabe, Reiji

    2009-01-01

    The aim of this study is to determine whether the body surface index replaces the Haller index in order to evaluate the severity of pectus excavatum. Sixty-nine cases with a diagnosis of pectus excavatum between August 2001 and January 2008 were prospectively enrolled in the study. The anterior-posterior chest dimension, lateral chest dimension and depth of the most profound depressed area were measured. The body surface index was expressed as an equation A/(B-C), where A was the transverse diameter, B was the anterior-posterior diameter and C was the depth of the depressed area. This index was compared to the Haller index in each patient. In this study, each value was measured on a chest CT scan. The body surface index significantly correlated with the Haller index, in which the coefficient of correlation was 0.879 (p<0.01). The body surface index corresponding to Haller index 3.25 seems to be 1.67. We could find the body surface index corresponding to each Haller index. We propose that an evaluation of pectus excavatum by means of body surface measurements is clinically useful, and may replace the Haller index. The body surface index may allow us to assess the severity of pectus excavatum without a CT scan, resulting in a reduction of radiation exposure for children with pectus excavatum. It's possible to evaluate the body surface index by using a relational table between this index and the Hailer index. We'll evaluate the body surface index by using values measured on a real body. (author)

  12. Optimization of digital chest radiography using computer modeling and voxels phantoms

    International Nuclear Information System (INIS)

    Correa, S.C.A.; Souza, E.M.; Silva, A.X.; Lopes, R.T.

    2009-01-01

    The purpose of this work is to use the Monte Carlo code MCNPX and the Female Adult voxel (FAX) and Male Adult voxel (MAX) phantoms to investigate how the dose and image quality in digital chest radiography vary with tube voltage (70-150 kV), anti-scatter methods (grid and air gap) and gender of the patient. The effective dose was calculated by ICRP60 and image quality was quantified by calculating the signal-difference-to-noise ratio for pathological details (calcifications) positioned at different locations in the anatomy. Calculated quantities were normalized to a fixed value of air kerma (5 μGy) at the automatic exposure control chambers. The results obtained in this work show that the air gap technique and lower tube voltages provide an increase in the digital image quality. Furthermore, this study has also shown that the detection of pathological details vary with the gender of the patient. (author)

  13. Utility of Objective Chest Tube Management After Pulmonary Resection Using a Digital Drainage System.

    Science.gov (United States)

    Takamochi, Kazuya; Imashimizu, Kota; Fukui, Mariko; Maeyashiki, Tatsuo; Suzuki, Mikiko; Ueda, Takuya; Matsuzawa, Hironori; Hirayama, Shunki; Matsunaga, Takeshi; Oh, Shiaki; Suzuki, Kenji

    2017-07-01

    We sought to evaluate the clinical utility of chest tube management after pulmonary resection based on objective digital monitoring of pleural pressure and digital surveillance for air leaks. We prospectively recorded the perioperative data of 308 patients who underwent pulmonary resection between December 2013 and January 2016. We used information from a digital monitoring thoracic drainage system to measure peak air leakage during the first 24 hours after the operation, patterns of air leakage over the first 72 hours, and patterns of pleural pressure changes until the chest tubes were removed. There were 240 patients with lung cancer and 68 patients with other diseases. The operations included 49 wedge resections, 58 segmentectomies, and 201 lobectomies. A postoperative air leak was observed in 61 patients (20%). A prolonged air leak exceeding 20 mL/min lasting 5 days or more was observed in 18 patients (5.8%). Multivariate analysis of various perioperative factors showed forced expiratory volume in 1 second below 70%, patterns of air leakage, defined as exacerbating and remitting or without a trend toward improvement, and peak air leakage of 100 mL/min or more were significant positive predictors of prolonged air leak. Fluctuations in pleural pressure occurred just after the air leakage rate decreased to less than 20 mL/min. Digital monitoring of peak air leakage and patterns of air leakage were useful for predicting prolonged air leak after pulmonary resection. Information on the disappearance of air leak could be derived from the change in the rate of air leakage and from the increase in fluctuation of pleural pressure. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  14. Noncontrast chest computed tomography immediately after transarterial chemoembolization in patients with hepatocellular carcinoma: Clinical benefits and effect of radiation reduction on image quality in low-dose scanning

    International Nuclear Information System (INIS)

    Choi, Joon-Il; Kim, Hyun Beom; Kim, Min Ju; Lee, Jong Seok; Koh, Young Whan; An, Sang Bu; Ko, Heung-kyu; Park, Joong-Won

    2011-01-01

    Purpose: To evaluate the clinical benefits of noncontrast chest computed tomography (CT) immediately after transarterial chemoembolization in patients with hepatocellular carcinoma and to assess the effect of radiation reduction on image quality in low-dose scanning. Materials and methods: From June to October 2010, we performed standard-dose, noncontrast chest CTs immediately after transarterial chemoembolization in 160 patients and low-dose CTs in 88 patients. We reviewed the entire noncontrast chest CTs and follow-up CTs to reveal the clinical benefits of CT evaluation immediately after transarterial chemoembolization. Using two independent readers, we also retrospectively evaluated the radiation dose and image quality in terms of the image noise, contrast between the liver parenchyma and iodized oil and diagnostic acceptability for the evaluation of treatment response after transarterial chemoembolization. Results: In 5.2% of the patients, additional treatment was performed immediately after the interpretation of the noncontrast chest CT, and additional pulmonary lesions were found in 8.5% of the patients. The measured mean dose-length product for the low-dose scanning was 18.4% of that of the standard-dose scanning. The image noise was significantly higher with the low-dose scanning (p < 0.001). However, all of the low-dose CT scans were diagnostically acceptable, and the mean scores for the subjective assessments of the contrast and diagnostic acceptability showed no significant differences for either reader. Conclusion: A noncontrast chest CT immediately after transarterial chemoembolization has some clinical benefits for immediate decision making and detecting pulmonary lesions. Low-dose, noncontrast chest CTs immediately after transarterial chemoembolization consistently provide diagnostically acceptable images and information on treatment response in patients who have undergone transarterial chemoembolization.

  15. Screening for lung cancer with digital chest radiography: sensitivity and number of secondary work-up CT examinations

    NARCIS (Netherlands)

    de Hoop, Bartjan; Schaefer-Prokop, Cornelia; Gietema, Hester A.; de Jong, Pim A.; van Ginneken, Bram; van Klaveren, Rob J.; Prokop, Mathias

    2010-01-01

    To estimate the performance of digital chest radiography for detection of lung cancer. The study had ethics committee approval, and a nested case-control design was used and included 55 patients with lung cancer detected at computed tomography (CT) and confirmed with histologic examination and a

  16. Optical scanning holography based on compressive sensing using a digital micro-mirror device

    Science.gov (United States)

    A-qian, Sun; Ding-fu, Zhou; Sheng, Yuan; You-jun, Hu; Peng, Zhang; Jian-ming, Yue; xin, Zhou

    2017-02-01

    Optical scanning holography (OSH) is a distinct digital holography technique, which uses a single two-dimensional (2D) scanning process to record the hologram of a three-dimensional (3D) object. Usually, these 2D scanning processes are in the form of mechanical scanning, and the quality of recorded hologram may be affected due to the limitation of mechanical scanning accuracy and unavoidable vibration of stepper motor's start-stop. In this paper, we propose a new framework, which replaces the 2D mechanical scanning mirrors with a Digital Micro-mirror Device (DMD) to modulate the scanning light field, and we call it OSH based on Compressive Sensing (CS) using a digital micro-mirror device (CS-OSH). CS-OSH can reconstruct the hologram of an object through the use of compressive sensing theory, and then restore the image of object itself. Numerical simulation results confirm this new type OSH can get a reconstructed image with favorable visual quality even under the condition of a low sample rate.

  17. Occult pneumomediastinum in blunt chest trauma: clinical significance.

    Science.gov (United States)

    Rezende-Neto, J B; Hoffmann, J; Al Mahroos, M; Tien, H; Hsee, L C; Spencer Netto, F; Speers, V; Rizoli, S B

    2010-01-01

    Thoracic injuries are potentially responsible for 25% of all trauma deaths. Chest X-ray is commonly used to screen patients with chest injury. However, the use of computed tomography (CT) scan for primary screening is increasing, particularly for blunt trauma. CT scans are more sensitive than chest X-ray in detecting intra-thoracic abnormalities such as pneumothoraces and pneumomediastinums. Pneumomediastinum detected by chest X-ray or "overt pneumomediastinum", raises the concern of possible aerodigestive tract injuries. In contrast, there is scarce information on the clinical significance of pneumomediastinum diagnosed by CT scan only or "occult pneumomediastinum". Therefore we investigated the clinical consequences of occult pneumomediastinum in our blunt trauma population. A 2-year retrospective chart review of all blunt chest trauma patients with initial chest CT scan admitted to a level I trauma centre. Data extracted from the medical records include; demographics, occult, overt, or no pneumomediastinum, the presence of intra-thoracic aerodigestive tract injuries (trachea, bronchus, and/or esophagus), mechanism and severity of injury, endotracheal intubation, chest thoracostomy, operations and radiological reports by an attending radiologist. All patients with intra-thoracic aerodigestive tract injuries from 1994 to 2004 were also investigated. Of 897 patients who met the inclusion criteria 839 (93.5%) had no pneumomediastinum. Five patients (0.6%) had overt pneumomediastinum and 53 patients (5.9%) had occult pneumomediastinum. Patients with occult pneumomediastinum had significantly higher ISS and AIS chest (pchest thoracostomy tube was more common (ppneumothorax. None of the patients with occult pneumomediastinum had aerodigestive tract injuries (95%CI 0-0.06). Follow up CT scan of patients with occult pneumomediastinum showed complete resolution in all cases, in average 3 h after the initial exam. Occult pneumomediastinum occurred in approximately 6% of

  18. Optimization of protocol in computerized tomography scans of children's chest

    International Nuclear Information System (INIS)

    Lima, L.T.A.; Roberto, W. dos S.; Mourão Filho, A.P.

    2017-01-01

    The diffusion of the Computed Tomography (CT) has promoted a significant increase of the dose absorbed by patients. This technique showed a high growth as a result of clinical cases of medical emergencies, neoplasia and pediatric traumas. The risks of stochastic effects are greater for children due to tissue radiosensitivity coupled with longer life expectancy. In the present work, a cylindrical simulator was used, representing an adult thorax made of polymethylmethacrylate (PMMA), and a second simulator object of the same material was developed in oblong format including the axillary regions, based on the dimensions of the patient's thorax one-year-old pediatric population. The scanning on the adult simulator object was performed with the chest radiograph diagnostic service with a 120 kV supply voltage and in the pediatric simulator object were performed optimized scans with 120, 100 and 80 kV. This work compared the dose variation absorbed by the pediatric simulator object mentioned by the change in the tube supply voltage

  19. Evaluation of the distortions of the digital chest image caused by the data compression

    International Nuclear Information System (INIS)

    Ando, Yutaka; Kunieda, Etsuo; Ogawa, Koichi; Tukamoto, Nobuhiro; Hashimoto, Shozo; Aoki, Makoto; Kurotani, Kenichi.

    1988-01-01

    The image data compression methods using orthogonal transforms (Discrete cosine transform, Discrete fourier transform, Hadamard transform, Haar transform, Slant transform) were analyzed. From the points of the error and the speed of the data conversion, the discrete cosine transform method (DCT) is superior to the other methods. The block quantization by the DCT for the digital chest image was used. The quality of data compressed and reconstructed images by the score analysis and the ROC curve analysis was examined. The chest image with the esophageal cancer and metastatic lung tumors was evaluated at the 17 checkpoints (the tumor, the vascular markings, the border of the heart and ribs, the mediastinal structures and et al). By our score analysis, the satisfactory ratio of the data compression is 1/5 and 1/10. The ROC analysis using normal chest images superimposed by the artificial coin lesions was made. The ROC curve of the 1/5 compressed ratio is almost as same as the original one. To summarize our study, the image data compression method using the DCT is thought to be useful for the clinical use and the 1/5 compression ratio is a tolerable ratio. (author)

  20. Evaluation of the distortions of the digital chest image caused by the data compression

    Energy Technology Data Exchange (ETDEWEB)

    Ando, Yutaka; Kunieda, Etsuo; Ogawa, Koichi; Tukamoto, Nobuhiro; Hashimoto, Shozo; Aoki, Makoto; Kurotani, Kenichi

    1988-08-01

    The image data compression methods using orthogonal transforms (Discrete cosine transform, Discrete fourier transform, Hadamard transform, Haar transform, Slant transform) were analyzed. From the points of the error and the speed of the data conversion, the discrete cosine transform method (DCT) is superior to the other methods. The block quantization by the DCT for the digital chest image was used. The quality of data compressed and reconstructed images by the score analysis and the ROC curve analysis was examined. The chest image with the esophageal cancer and metastatic lung tumors was evaluated at the 17 checkpoints (the tumor, the vascular markings, the border of the heart and ribs, the mediastinal structures and et al). By our score analysis, the satisfactory ratio of the data compression is 1/5 and 1/10. The ROC analysis using normal chest images superimposed by the artificial coin lesions was made. The ROC curve of the 1/5 compressed ratio is almost as same as the original one. To summarize our study, the image data compression method using the DCT is thought to be useful for the clinical use and the 1/5 compression ratio is a tolerable ratio.

  1. Marginal and internal fit of zirconia copings obtained using different digital scanning methods

    Directory of Open Access Journals (Sweden)

    Lorena Oliveira PEDROCHE

    Full Text Available Abstract The objective of this study was to evaluate the marginal and internal fit of zirconia copings obtained with different digital scanning methods. A human mandibular first molar was set in a typodont with its adjacent and antagonist teeth and prepared for an all-ceramic crown. Digital impressions were made using an intraoral scanner (3Shape. Polyvinyl siloxane impressions and Type IV gypsum models were also obtained and scanned with a benchtop laboratory scanner (3Shape D700. Ten zirconia copings were fabricated for each group using CAD-CAM technology. The marginal and internal fit of the zirconia copings was assessed by the silicone replica technique. Four sections of each replica were obtained, and each section was evaluated at four points: marginal gap (MG, axial wall (AW, axio-occlusal edge (AO and centro-occlusal wall (CO, using an image analyzing software. The data were submitted to one-way ANOVA and Tukey’s test (α = 0.05. They showed statistically significant differences for MG, AO and CO. Regarding MG, intraoral scanning showed lower gap values, whereas gypsum model scanning showed higher gap values. Regarding AO and CO, intraoral digital scanning showed lower gap values. Polyvinyl siloxane impression scanning and gypsum model scanning showed higher gap values and were statistically similar. It can be concluded that intraoral digital scanning provided a lower mean gap value, in comparison with conventional impressions and gypsum casts scanned with a standard benchtop laboratory scanner.

  2. Chest radiology

    International Nuclear Information System (INIS)

    Reed, J.C.

    1990-01-01

    This book is a reference in plain chest film diagnosis provides a thorough background in the differential diagnosis of 22 of the most common radiologic patterns of chest disease. Each chapter is introduced with problem cases and a set of questions, followed by a tabular listing of the appropriate differential considerations. The book emphasizes plain films, CT and some MR scans are integrated to demonstrate how these modalities enhance the work of a case

  3. A feasibility study for anatomical noise reduction in dual-energy chest digital tomosynthesis

    Science.gov (United States)

    Lee, D.; Kim, Y.-s.; Choi, S.; Lee, H.; Choi, S.; Kim, H.-J.

    2016-01-01

    Lung cancer is the leading cause of cancer death worldwide. Thus, early diagnosis is of considerable importance. For early screening of lung cancer, computed tomography (CT) has been used as the gold standard. Chest digital tomosynthesis (CDT) is a recently introduced modality for lung cancer screening with a relatively low radiation dose compared to CT. The dual energy material decomposition method has been proposed for better detection of pulmonary nodules by means of reducing anatomical noise. In this study, the possibility of material decomposition in CDT was tested by both a simulation study and an experimental study using a CDT prototype. The Geant4 application for tomographic emission (GATE) v6 and tungsten anode spectral model using interpolating polynomials (TASMIP) codes were used for the simulation study to create simulated phantom shapes consisting of five inner cylinders filled with different densities of bone and airequivalent materials. Furthermore, the CDT prototype system and human phantom chest were used for the experimental study. CDT scan in both the simulation and experimental studies was performed with linear movement and 21 projection images were obtained over a 30 degree angular range with a 1.5 degree angular interval. To obtain materialselective images, a projectionbased energy subtraction technique was applied to high and low energy images. The resultant simulation images showed that dual-energy reconstruction could achieve an approximately 32% higher contrast to noise ratio (CNR) in images and the difference in CNR value according to bone density was significant compared to single energy CDT. Additionally, image artifacts were effectively corrected in dual energy CDT simulation studies. Likewise the experimental study with dual energy produced clear images of lung fields and bone structure by removing unnecessary anatomical structures. Dual energy tomosynthesis is a new technique; therefore, there is little guidance regarding its

  4. A feasibility study for anatomical noise reduction in dual-energy chest digital tomosynthesis

    International Nuclear Information System (INIS)

    Lee, D.; Choi, S.; Kim, H.-J.; Kim, Y.-S.; Choi, S.; Lee, H.

    2016-01-01

    Lung cancer is the leading cause of cancer death worldwide. Thus, early diagnosis is of considerable importance. For early screening of lung cancer, computed tomography (CT) has been used as the gold standard. Chest digital tomosynthesis (CDT) is a recently introduced modality for lung cancer screening with a relatively low radiation dose compared to CT. The dual energy material decomposition method has been proposed for better detection of pulmonary nodules by means of reducing anatomical noise. In this study, the possibility of material decomposition in CDT was tested by both a simulation study and an experimental study using a CDT prototype. The Geant4 application for tomographic emission (GATE) v6 and tungsten anode spectral model using interpolating polynomials (TASMIP) codes were used for the simulation study to create simulated phantom shapes consisting of five inner cylinders filled with different densities of bone and airequivalent materials. Furthermore, the CDT prototype system and human phantom chest were used for the experimental study. CDT scan in both the simulation and experimental studies was performed with linear movement and 21 projection images were obtained over a 30 degree angular range with a 1.5 degree angular interval. To obtain materialselective images, a projectionbased energy subtraction technique was applied to high and low energy images. The resultant simulation images showed that dual-energy reconstruction could achieve an approximately 32% higher contrast to noise ratio (CNR) in images and the difference in CNR value according to bone density was significant compared to single energy CDT. Additionally, image artifacts were effectively corrected in dual energy CDT simulation studies. Likewise the experimental study with dual energy produced clear images of lung fields and bone structure by removing unnecessary anatomical structures. Dual energy tomosynthesis is a new technique; therefore, there is little guidance regarding its

  5. Optimization of scanning strategy of digital Shack-Hartmann wavefront sensing.

    Science.gov (United States)

    Guo, Wenjiang; Zhao, Liping; Li, Xiang; Chen, I-Ming

    2012-01-01

    In the traditional Shack-Hartmann wavefront sensing (SHWS) system, a lenslet array with a bigger configuration is desired to achieve a higher lateral resolution. However, practical implementation limits the configuration and this parameter is contradicted with the measurement range. We have proposed a digital scanning technique by making use of the high flexibility of a spatial light modulator to sample the reflected wavefront [X. Li, L. P. Zhao, Z. P. Fang, and C. S. Tan, "Improve lateral resolution in wavefront sensing with digital scanning technique," in Asia-Pacific Conference of Transducers and Micro-Nano Technology (2006)]. The lenslet array pattern is programmed to laterally scan the whole aperture. In this paper, the methodology to optimize the scanning step for the purpose of form measurement is proposed. The correctness and effectiveness are demonstrated in numerical simulation and experimental investigation. © 2012 Optical Society of America

  6. Measurement of residual stress by using focused ion beam and digital image correlation method in thin-sized wires used for steel cords

    International Nuclear Information System (INIS)

    Yang, Y S; Park, C G; Bae, J G

    2008-01-01

    Residual stress in the axial direction of the steel wires has been measured by using a method based on the combination of the focused ion beam (FIB) milling and digital image correlation software. That is, the residual stress was calculated from the measured displacement field before and after the introduction of a slot along the steel wires. The displacement was obtained by the digital correlation analysis of high-resolution scanning electron micrographs, while the slot was introduced by FIB milling with low energy beam. The fitting of the experimental results to an analytical model with the independent Young's modulus determined allows us to find the residual stress. The complete experimental procedures are described and its feasibilities are also evaluated for the thin-sized steel wires

  7. Digital chest radiography: collimation and dose reduction

    DEFF Research Database (Denmark)

    Debess, Jeanne; Johnsen, Karen Kirstine; Vejle-Sørensen, Jens Kristian

    ,3 mAs and SID SID of 180 centimetres using a phantom and lithium fluoride thermo luminescence dosimeter (TLD). Dose to risk organs mamma, thyroid and colon are measured at different collimations with one-centimetre steps. TLD results are used to estimate dose reduction for different collimations...... at the conference. Conclusion: Collimation improvement in basic chest radiography can reduce the radiation to female patients at chest x-ray examinations....

  8. A design procedure for a slotted waveguide with probe-fed slots radiating into plasma

    International Nuclear Information System (INIS)

    Colborn, J.A.

    1989-11-01

    A design procedure is developed for slotted-waveguide antennas with probe-fed slots. Radiation into a gyrotropic, plane-stratified medium is considered, nonzero waveguide wall thickness is assumed, and noncosinusoidal slot fields and arbitrary slot length up to about one free-space wavelength are allowed. External mutual coupling is taken into account by matching the tangential fields at the antenna surface. The particular case of longitudinal slots in the broad face of rectangular guide is analyzed. The motivation for this work is the design of such radiators for plasma heating and current-drive on thermonuclear fusion experiments, but some of the analysis is applicable to the probeless slotted waveguide used for avionics and communications. 20 refs., 5 figs

  9. Parallel-scanning tomosynthesis using a slot scanning technique: Fixed-focus reconstruction and the resulting image quality

    International Nuclear Information System (INIS)

    Shibata, Koichi; Notohara, Daisuke; Sakai, Takihito

    2014-01-01

    Purpose: Parallel-scanning tomosynthesis (PS-TS) is a novel technique that fuses the slot scanning technique and the conventional tomosynthesis (TS) technique. This approach allows one to obtain long-view tomosynthesis images in addition to normally sized tomosynthesis images, even when using a system that has no linear tomographic scanning function. The reconstruction technique and an evaluation of the resulting image quality for PS-TS are described in this paper. Methods: The PS-TS image-reconstruction technique consists of several steps (1) the projection images are divided into strips, (2) the strips are stitched together to construct images corresponding to the reconstruction plane, (3) the stitched images are filtered, and (4) the filtered stitched images are back-projected. In the case of PS-TS using the fixed-focus reconstruction method (PS-TS-F), one set of stitched images is used for the reconstruction planes at all heights, thus avoiding the necessity of repeating steps (1)–(3). A physical evaluation of the image quality of PS-TS-F compared with that of the conventional linear TS was performed using a R/F table (Sonialvision safire, Shimadzu Corp., Kyoto, Japan). The tomographic plane with the best theoretical spatial resolution (the in-focus plane, IFP) was set at a height of 100 mm from the table top by adjusting the reconstruction program. First, the spatial frequency response was evaluated at heights of −100, −50, 0, 50, 100, and 150 mm from the IFP using the edge of a 0.3-mm-thick copper plate. Second, the spatial resolution at each height was visually evaluated using an x-ray test pattern (Model No. 38, PTW Freiburg, Germany). Third, the slice sensitivity at each height was evaluated via the wire method using a 0.1-mm-diameter tungsten wire. Phantom studies using a knee phantom and a whole-body phantom were also performed. Results: The spatial frequency response of PS-TS-F yielded the best results at the IFP and degraded slightly as the

  10. Parallel-scanning tomosynthesis using a slot scanning technique: fixed-focus reconstruction and the resulting image quality.

    Science.gov (United States)

    Shibata, Koichi; Notohara, Daisuke; Sakai, Takihito

    2014-11-01

    Parallel-scanning tomosynthesis (PS-TS) is a novel technique that fuses the slot scanning technique and the conventional tomosynthesis (TS) technique. This approach allows one to obtain long-view tomosynthesis images in addition to normally sized tomosynthesis images, even when using a system that has no linear tomographic scanning function. The reconstruction technique and an evaluation of the resulting image quality for PS-TS are described in this paper. The PS-TS image-reconstruction technique consists of several steps (1) the projection images are divided into strips, (2) the strips are stitched together to construct images corresponding to the reconstruction plane, (3) the stitched images are filtered, and (4) the filtered stitched images are back-projected. In the case of PS-TS using the fixed-focus reconstruction method (PS-TS-F), one set of stitched images is used for the reconstruction planes at all heights, thus avoiding the necessity of repeating steps (1)-(3). A physical evaluation of the image quality of PS-TS-F compared with that of the conventional linear TS was performed using a R/F table (Sonialvision safire, Shimadzu Corp., Kyoto, Japan). The tomographic plane with the best theoretical spatial resolution (the in-focus plane, IFP) was set at a height of 100 mm from the table top by adjusting the reconstruction program. First, the spatial frequency response was evaluated at heights of -100, -50, 0, 50, 100, and 150 mm from the IFP using the edge of a 0.3-mm-thick copper plate. Second, the spatial resolution at each height was visually evaluated using an x-ray test pattern (Model No. 38, PTW Freiburg, Germany). Third, the slice sensitivity at each height was evaluated via the wire method using a 0.1-mm-diameter tungsten wire. Phantom studies using a knee phantom and a whole-body phantom were also performed. The spatial frequency response of PS-TS-F yielded the best results at the IFP and degraded slightly as the distance from the IFP increased. A

  11. Parallel-scanning tomosynthesis using a slot scanning technique: Fixed-focus reconstruction and the resulting image quality

    Energy Technology Data Exchange (ETDEWEB)

    Shibata, Koichi, E-mail: shibatak@suzuka-u.ac.jp [Department of Radiological Technology, Faculty of Health Science, Suzuka University of Medical Science 1001-1, Kishioka-cho, Suzuka 510-0293 (Japan); Notohara, Daisuke; Sakai, Takihito [R and D Department, Medical Systems Division, Shimadzu Corporation 1, Nishinokyo-Kuwabara-cho, Nakagyo-ku, Kyoto 604-8511 (Japan)

    2014-11-01

    Purpose: Parallel-scanning tomosynthesis (PS-TS) is a novel technique that fuses the slot scanning technique and the conventional tomosynthesis (TS) technique. This approach allows one to obtain long-view tomosynthesis images in addition to normally sized tomosynthesis images, even when using a system that has no linear tomographic scanning function. The reconstruction technique and an evaluation of the resulting image quality for PS-TS are described in this paper. Methods: The PS-TS image-reconstruction technique consists of several steps (1) the projection images are divided into strips, (2) the strips are stitched together to construct images corresponding to the reconstruction plane, (3) the stitched images are filtered, and (4) the filtered stitched images are back-projected. In the case of PS-TS using the fixed-focus reconstruction method (PS-TS-F), one set of stitched images is used for the reconstruction planes at all heights, thus avoiding the necessity of repeating steps (1)–(3). A physical evaluation of the image quality of PS-TS-F compared with that of the conventional linear TS was performed using a R/F table (Sonialvision safire, Shimadzu Corp., Kyoto, Japan). The tomographic plane with the best theoretical spatial resolution (the in-focus plane, IFP) was set at a height of 100 mm from the table top by adjusting the reconstruction program. First, the spatial frequency response was evaluated at heights of −100, −50, 0, 50, 100, and 150 mm from the IFP using the edge of a 0.3-mm-thick copper plate. Second, the spatial resolution at each height was visually evaluated using an x-ray test pattern (Model No. 38, PTW Freiburg, Germany). Third, the slice sensitivity at each height was evaluated via the wire method using a 0.1-mm-diameter tungsten wire. Phantom studies using a knee phantom and a whole-body phantom were also performed. Results: The spatial frequency response of PS-TS-F yielded the best results at the IFP and degraded slightly as the

  12. Accuracy of a computed tomography scanning procedure to manufacture digital models.

    NARCIS (Netherlands)

    Darroudi, A.M.; Kuijpers-Jagtman, A.M.; Ongkosuwito, E.M.; Suttorp, C.M.; Bronkhorst, E.M.; Breuning, K.H.

    2017-01-01

    INTRODUCTION: Accurate articulation of the digital dental casts is crucial in orthodontic diagnosis and treatment planning. We aimed to determine the accuracy of manufacturing digital dental casts from computed tomography scanning of plaster casts regarding linear dimensions and interarch

  13. Digital acquisition and processing of electron micrographs using a scanning transmission electron microscope

    International Nuclear Information System (INIS)

    Engel, A.; Christen, F.; Michel, B.

    1981-01-01

    A digital acquisition system that collects multichannel information from a scanning transmission electron microscope (STEM) and its application are described. The hardware comprises (i) single electron counting detectors, (ii) a digital scan generator, (iii) a digital multi-channel on-line processor, (iv) an interface to a minicomputer, and (v) a display system. Experimental results characterizing these components are presented, and their performance is discussed. The software includes assembler coded programs for dynamic file maintenance and fast acquisition of image data, a display driver, and FORTRAN coded application programs. The usefulness of digitized STEM is illustrated by a variety of biological applications. (orig.)

  14. Development and image quality assessment of a contrast-enhancement algorithm for display of digital chest radiographs

    International Nuclear Information System (INIS)

    Rehm, K.

    1992-01-01

    This dissertation presents a contrast-enhancement algorithm Artifact-Suppressed Adaptive Histogram Equalization (ASAHE). This algorithm was developed as part of a larger effort to replace the film radiographs currently used in radiology departments with digital images. Among the expected benefits of digital radiology are improved image management and greater diagnostic accuracy. Film radiographs record X-ray transmission data at high spatial resolution, and a wide dynamic range of signal. Current digital radiography systems record an image at reduced spatial resolution and with coarse sampling of the available dynamic range. These reductions have a negative impact on diagnostic accuracy. The contrast-enhancement algorithm presented in this dissertation is designed to boost diagnostic accuracy of radiologists using digital images. The ASAHE algorithm is an extension of an earlier technique called Adaptive Histogram Equalization (AHE). The AHE algorithm is unsuitable for chest radiographs because it over-enhances noise, and introduces boundary artifacts. The modifications incorporated in ASAHE suppress the artifacts and allow processing of chest radiographs. This dissertation describes the psychophysical methods used to evaluate the effects of processing algorithms on human observer performance. An experiment conducted with anthropomorphic phantoms and simulated nodules showed the ASAHE algorithm to be superior for human detection of nodules when compared to a computed radiography system's algorithm that is in current use. An experiment conducted using clinical images demonstrating pneumothoraces (partial lung collapse) indicated no difference in human observer accuracy when ASAHE images were compared to computed radiography images, but greater ease of diagnosis when ASAHE images were used. These results provide evidence to suggest that Artifact-Suppressed Adaptive Histogram Equalization can be effective in increasing diagnostic accuracy and efficiency

  15. Digital radiology. Clinical experience

    Energy Technology Data Exchange (ETDEWEB)

    Stacul, F; Smathers, R L

    1985-01-01

    The authors report the experience achieved ad the Stanford University (USA) with a digital radiography system which allows the digitization of the film and of the images collected with photostimulable phophors. The phophor is essentially an intensifying screen where a latent image is stored after exposure to X-rays and is extracted by a laser scanning. The images collected with the digitized film and with the phophor (chest, breast, bone) have been analyzed. The digitized film offers potential diagnostic advantages over the conventional film, because of the contrast manipulation and many other processing options. The possibility to recover the information of overexposed films appears very attractive. The photostimulable phophors allow to get good quality images, with a consistent reduction of dose and costs. These plates offer the possibility, in the next future, to replace the conventional screen-film systems.

  16. Comparison of the image quality of digital radiography system and film screen system - Radiologist' rating of the visibility of normal anatomic - Structures in chest PA, Skull radiograph and K. U. B

    International Nuclear Information System (INIS)

    Song, Kounn Sik; Kim, Young Goo; Lee, Jong Beum; Kim, Kun Sang

    1987-01-01

    Digital image acquisition and display is widely used in computed tomography, ultrasonography, digital subtraction angiography, nuclear medicine and magnetic resonance image. But most of the radiological examinations performed in radiology department are made by using conventional system. The development of the digital radiography system is essential if totally digitized radiology department is desired. The advantages of digitizing the radiographic information are usually discussed in terms of PACS (picture archiving and communication system), furthermore there are many other advantages such as contrast modification, spatial filtering subtraction and superimposition of the images through the image processing by computer. Currently several approaches are under development or in clinical use, the most promising approach is the use of imaging plate composed of photostimulate phosphors such as barium fluorohalide crystal read with a He-Ne laser to produce digital radiographic images. Another promising approach is scan projection radiography. The authors performed the clinical study of comparing the image qualities of digital radiography system using scanning laser luminescence (FCR) and conventional film-screen system in chest PA, skull radiography and K. U. B. in terms of the visibility of the normal anatomic structure rating those (qualities) on a scale of 0 to 3 and obtained the following results. Normal contrast digital images are comparable to conventional film-screen images, but the images of high frequency enhancement is far superior to conventional film-screen especially in peripherally located structures such as skin, subcutaneous fat, musculoskeletal systems, nasal bone, inner and outer table of the skull including the diploic space, paranasal sinuses, nasopharynx and larynx, trachea and main bronchi, mediastinal structures, retrocardiac and subphrenic vascular markings. Another promising aspects of digital radiography system is its wide exposure latitude and

  17. Comparison study of noise reduction algorithms in dual energy chest digital tomosynthesis

    Science.gov (United States)

    Lee, D.; Kim, Y.-S.; Choi, S.; Lee, H.; Choi, S.; Kim, H.-J.

    2018-04-01

    Dual energy chest digital tomosynthesis (CDT) is a recently developed medical technique that takes advantage of both tomosynthesis and dual energy X-ray images. However, quantum noise, which occurs in dual energy X-ray images, strongly interferes with diagnosis in various clinical situations. Therefore, noise reduction is necessary in dual energy CDT. In this study, noise-compensating algorithms, including a simple smoothing of high-energy images (SSH) and anti-correlated noise reduction (ACNR), were evaluated in a CDT system. We used a newly developed prototype CDT system and anthropomorphic chest phantom for experimental studies. The resulting images demonstrated that dual energy CDT can selectively image anatomical structures, such as bone and soft tissue. Among the resulting images, those acquired with ACNR showed the best image quality. Both coefficient of variation and contrast to noise ratio (CNR) were the highest in ACNR among the three different dual energy techniques, and the CNR of bone was significantly improved compared to the reconstructed images acquired at a single energy. This study demonstrated the clinical value of dual energy CDT and quantitatively showed that ACNR is the most suitable among the three developed dual energy techniques, including standard log subtraction, SSH, and ACNR.

  18. Efficacy of daily bedside chest radiography as visualized by digital luminescence radiography

    International Nuclear Information System (INIS)

    Kirchner, J.; Stueckle, C.A.; Schilling, E.M.; Peters, J.

    2001-01-01

    To determine the diagnostic impact of daily bedside chest radiography in comparison with digital luminescence technique (DLR; storage phosphor radiography) and conventional film screen radiography, a prospective randomized study was completed in 210 mechanically ventilated patients with a total of 420 analysed radiographs. The patients were allocated to two groups: 150 patients underwent DLR, and 60 patients underwent conventional film screen radiography. Radiological analysis was performed consensually and therapeutic efficacy was assessed by the clinicians. There was no statistical significant difference between the frequency of abnormal findings seen on DLR and conventional film screen radiography. In total, 448 abnormal findings were present in 249 of 300 DLR and 97 of 120 conventional film screen radiographs. The most common findings were signs of overhydration (41 %), pleural effusion (31%), partial collapse of the lung (11%) and pneumothorax (2%). One hundred and twenty-three of 448 (27%) of these abnormal findings were thought to have a considerable impact on patient management. The high rate of abnormal findings with significant impact on patient management suggests that the use of daily bedside chest radiography may be reasonable. Copyright (2001) Blackwell Science Pty Ltd

  19. Chest wall segmentation in automated 3D breast ultrasound scans.

    Science.gov (United States)

    Tan, Tao; Platel, Bram; Mann, Ritse M; Huisman, Henkjan; Karssemeijer, Nico

    2013-12-01

    In this paper, we present an automatic method to segment the chest wall in automated 3D breast ultrasound images. Determining the location of the chest wall in automated 3D breast ultrasound images is necessary in computer-aided detection systems to remove automatically detected cancer candidates beyond the chest wall and it can be of great help for inter- and intra-modal image registration. We show that the visible part of the chest wall in an automated 3D breast ultrasound image can be accurately modeled by a cylinder. We fit the surface of our cylinder model to a set of automatically detected rib-surface points. The detection of the rib-surface points is done by a classifier using features representing local image intensity patterns and presence of rib shadows. Due to attenuation of the ultrasound signal, a clear shadow is visible behind the ribs. Evaluation of our segmentation method is done by computing the distance of manually annotated rib points to the surface of the automatically detected chest wall. We examined the performance on images obtained with the two most common 3D breast ultrasound devices in the market. In a dataset of 142 images, the average mean distance of the annotated points to the segmented chest wall was 5.59 ± 3.08 mm. Copyright © 2012 Elsevier B.V. All rights reserved.

  20. Correlative analysis of longitudinal changes in bronchoalveolar lavage, 67Gallium scanning, serum angiotensin-converting enzyme activity, chest x-ray, and pulmonary function tests in pulmonary sarcoidosis

    International Nuclear Information System (INIS)

    Okada, Mitsuko; Takahashi, Hideki; Nukiwa, Toshihiro; Matsuoka, Rokuro; Furuse, Makoto; Kitamura, Satoshi; Kira, Shiro.

    1987-01-01

    Despite the relatively high cost and complicated procedures, Gallium-67 ( 67 Ga) scanning and bronchoalveolar lavage (BAL) are increasingly advocated as more sensitive indicators of disease activity in sarcoidosis than chest X-ray and serum angiotensin-converting enzyme activity (SACE). To evaluate the clinical usefulness of 67 Ga scanning and BAL, we followed 31 patients with pulmonary sarcoidosis, using these four parameters, at 9- to 24-month intervals over periods of 9 to 48 months. We obtained 68 complete evaluations. Close correlations were observed among chest X-ray, 67 Ga scanning, SACE, and the percent-age of lymphocytes in BAL fluid (p 67 Ga scanning and BAL are not necessarily indicated in the long-term management of pulmonary sarcoidosis. (author)

  1. Precision of guided scanning procedures for full-arch digital impressions in vivo.

    Science.gov (United States)

    Zimmermann, Moritz; Koller, Christina; Rumetsch, Moritz; Ender, Andreas; Mehl, Albert

    2017-11-01

    System-specific scanning strategies have been shown to influence the accuracy of full-arch digital impressions. Special guided scanning procedures have been implemented for specific intraoral scanning systems with special regard to the digital orthodontic workflow. The aim of this study was to evaluate the precision of guided scanning procedures compared to conventional impression techniques in vivo. Two intraoral scanning systems with implemented full-arch guided scanning procedures (Cerec Omnicam Ortho; Ormco Lythos) were included along with one conventional impression technique with irreversible hydrocolloid material (alginate). Full-arch impressions were taken three times each from 5 participants (n = 15). Impressions were then compared within the test groups using a point-to-surface distance method after best-fit model matching (OraCheck). Precision was calculated using the (90-10%)/2 quantile and statistical analysis with one-way repeated measures ANOVA and post hoc Bonferroni test was performed. The conventional impression technique with alginate showed the lowest precision for full-arch impressions with 162.2 ± 71.3 µm. Both guided scanning procedures performed statistically significantly better than the conventional impression technique (p Cerec Omnicam Ortho were 74.5 ± 39.2 µm and for group Ormco Lythos 91.4 ± 48.8 µm. The in vivo precision of guided scanning procedures exceeds conventional impression techniques with the irreversible hydrocolloid material alginate. Guided scanning procedures may be highly promising for clinical applications, especially for digital orthodontic workflows.

  2. Evaluation of the effect scan pattern has on the trueness and precision of six intraoral digital impression systems.

    Science.gov (United States)

    Mennito, Anthony S; Evans, Zachary P; Lauer, Abigail W; Patel, Ravi B; Ludlow, Mark E; Renne, Walter G

    2018-03-01

    Clinicians have been slow to adopt digital impression technologies due possibly to perceived technique sensitivities involved in data acquisition. This research has two aims: determine whether scan pattern and sequence affects the accuracy of the three-dimensional (3D) model created from this digital impression and to compare the 5 imaging systems with regards to their scanning accuracy for sextant impressions. Six digital intraoral impression systems were used to scan a typodont sextant with optical properties similar to natural teeth. The impressions were taken using five different scan patterns and the resulting digital models were overlayed on a master digital model to determine the accuracy of each scanner performing each scan pattern. Furthermore, regardless of scan pattern, each digital impression system was evaluated for accuracy to the other systems in this same manner. No differences of significance were noted in the accuracy of 3D models created using six distinct scan patterns with one exception involving the CEREC Omnicam. Planmeca Planscan was determined to be the truest scanner while 3Shape Trios was determined to be the most precise for sextant impression making. Scan pattern does not significantly affect the accuracy of the resulting digital model for sextant scanning. Companies who make digital impression systems often recommend a scan pattern specific for their system. However, every clinical scanning scenario is different and may require a different approach. Knowing how important scan pattern is with regards to accuracy would be helpful for guiding a growing number of practitioners who are utilizing this technology. © 2018 Wiley Periodicals, Inc.

  3. Radiation exposure from Chest CT: Issues and Strategies

    Science.gov (United States)

    Maher, Michael M.; Rizzo, Stefania; Kanarek, David; Shephard, Jo-Anne O.

    2004-01-01

    Concerns have been raised over alleged overuse of CT scanning and inappropriate selection of scanning methods, all of which expose patients to unnecessary radiation. Thus, it is important to identify clinical situations in which techniques with lower radiation dose such as plain radiography or no radiation such as MRI and occasionally ultrasonography can be chosen over CT scanning. This article proposes the arguments for radiation dose reduction in CT scanning of the chest and discusses recommended practices and studies that address means of reducing radiation exposure associated with CT scanning of the chest. PMID:15082885

  4. Three-Dimensional Digital Documentation of Heritage Sites Using Terrestrial Laser Scanning and Unmanned Aerial Vehicle Photogrammetry

    Science.gov (United States)

    Jo, Y. H.; Kim, J. Y.

    2017-08-01

    Three-dimensional digital documentation is an important technique for the maintenance and monitoring of cultural heritage sites. This study focuses on the three-dimensional digital documentation of the Magoksa Temple, Republic of Korea, using a combination of terrestrial laser scanning and unmanned aerial vehicle (UAV) photogrammetry. Terrestrial laser scanning mostly acquired the vertical geometry of the buildings. In addition, the digital orthoimage produced by UAV photogrammetry had higher horizontal data acquisition rate than that produced by terrestrial laser scanning. Thus, the scanning and UAV photogrammetry were merged by matching 20 corresponding points and an absolute coordinate system was established using seven ground control points. The final, complete threedimensional shape had perfect horizontal and vertical geometries. This study demonstrates the potential of integrating terrestrial laser scanning and UAV photogrammetry for three-dimensional digital documentation. This new technique is expected to contribute to the three-dimensional digital documentation and spatial analysis of cultural heritage sites.

  5. Integration of digital dental casts in cone-beam computed tomography scans

    NARCIS (Netherlands)

    Rangel, F.A.; Maal, T.J.J.; Berge, S.J.; Kuijpers-Jagtman, A.M.

    2012-01-01

    Cone-beam computed tomography (CBCT) is widely used in maxillofacial surgery. The CBCT image of the dental arches, however, is of insufficient quality to use in digital planning of orthognathic surgery. Several authors have described methods to integrate digital dental casts into CBCT scans, but all

  6. Diagnostic imaging costs before and after digital tomosynthesis implementation in patient management after detection of suspected thoracic lesions on chest radiography.

    Science.gov (United States)

    Quaia, Emilio; Grisi, Guido; Baratella, Elisa; Cuttin, Roberto; Poillucci, Gabriele; Kus, Sara; Cova, Maria Assunta

    2014-02-01

    To evaluate diagnostic imaging costs before and after DTS implementation in patients with suspected thoracic lesions on CXR. Four hundred sixty-five patients (263 male, 202 female; age, 72.47 ± 11.33 years) with suspected thoracic lesion(s) after CXR underwent DTS. Each patient underwent CT when a pulmonary non-calcified lesion was identified by DTS while CT was not performed when a benign pulmonary or extrapulmonary lesion or pseudolesion was identified. The average per-patient imaging cost was calculated by normalising the costs before and after DTS implementation. In 229/465 patients who underwent DTS after suspicious CXR, DTS showed 193 pulmonary lesions and 36 pleural lesions, while in the remaining 236/465 patients, lesions were ruled out as pseudolesions of CXR. Chest CT examination was performed in 127/465 (27 %) patients while in the remaining 338/465 patients (73 %) CXR doubtful findings were resolved by DTS. The average per-patient costs of CXR, DTS and CT were 15.15, 41.55 and 113.66. DTS allowed an annual cost saving of 8,090.2 considering unenhanced CT and 19,298.12 considering contrast-enhanced CT. Considering a DTS reimbursement rate of 62.7 the break even point corresponds to 479 DTS examinations. Per-patient diagnostic imaging costs decreased after DTS implementation in patients with suspected thoracic lesions. • Digital tomosynthesis improves the diagnostic accuracy and confidence in chest radiography • Digital tomosynthesis reduces the need for CT for a suspected pulmonary lesion • Digital tomosynthesis requires a dose level equivalent to that of around two chest radiographies • Digital tomosynthesis produces a significant per-patient saving in diagnostic imaging costs.

  7. Should the lateral chest radiograph be routinely performed?

    International Nuclear Information System (INIS)

    Osman, Fatuma; Williams, Imelda

    2014-01-01

    Background: The chest x-ray is one of the most common plain film radiographic examinations performed. Inclusion of the lateral chest radiograph varies internationally and nationally across radiology departments and states in Australia. Search strategy: A search strategy of the databases Cochrane Library, Ovid Medline/Medline, PubMed, Scopus and Science Direct was conducted. The results were restricted to those published between 1985 and 2013 and those published in English. The following search terms were used: ‘lateral chest’, ‘radiograph’, ‘digital radiography’, ‘chest x-ray’, ‘plain film radiography’, ‘ionising radiation’. The results were restricted to publications with these terms in the title, abstract and/or keywords. Main findings: There are few national or international guidelines pertaining to the inclusion of the lateral chest x-ray as routine. Primary concerns are the increased radiation dose associated with the additional chest view and reduction of medical imaging services cost. Modern digital imaging systems result in a lower radiation dose. The diagnostic yield of the lateral chest x-ray is highly dependent on the clinical indications of the patient. Further research into the routine inclusion of the lateral chest x-ray is recommended. Conclusion: Review of the literature suggests that the lateral chest radiograph should not be performed routinely unless clinically indicated

  8. Large pneumothorax in blunt chest trauma: Is a chest drain always necessary in stable patients? A case report.

    Science.gov (United States)

    Idris, Baig M; Hefny, Ashraf F

    2016-01-01

    Pneumothorax is the most common potentially life-threatening blunt chest injury. The management of pneumothorax depends upon the etiology, its size and hemodynamic stability of the patient. Most clinicians agree that chest drainage is essential for the management of traumatic large pneumothorax. Herein, we present a case of large pneumothorax in blunt chest trauma patient that resolved spontaneously without a chest drain. A 63- year- old man presented to the Emergency Department complaining of left lateral chest pain due to a fall on his chest at home. On examination, he was hemodynamically stable. An urgent chest X-ray showed evidence of left sided pneumothorax. CT scan of the chest showed pneumothorax of more than 30% of the left hemithorax (around 600ml of air) with multiple left ribs fracture. Patient refused tube thoracostomy and was admitted to surgical department for close observation. The patient was managed conservatively without chest tube insertion. A repeat CT scan of the chest has shown complete resolution of the pneumothorax. The clinical spectrum of pneumothorax varies from asymptomatic to life threatening tension pneumothorax. In stable patients, conservative management can be safe and effective for small pneumothorax. To the best of our knowledge, this is the second reported case in the English literature with large pneumothorax which resolved spontaneously without chest drain. Blunt traumatic large pneumothorax in a clinically stable patient can be managed conservatively. Current recommendations for tube placement may need to be reevaluated. This may reduce morbidity associated with chest tube thoracostomy. Copyright © 2016 The Authors. Published by Elsevier Ltd.. All rights reserved.

  9. Fine Metal Mask 3-Dimensional Measurement by using Scanning Digital Holographic Microscope

    Science.gov (United States)

    Shin, Sanghoon; Yu, Younghun

    2018-04-01

    For three-dimensional microscopy, fast and high axial resolution are very important. Extending the depth of field for digital holographic is necessary for three-dimensional measurements of thick samples. We propose an optical sectioning method for optical scanning digital holography that is performed in the frequency domain by spatial filtering of a reconstructed amplitude image. We established a scanning dual-wavelength off-axis digital holographic microscope to measure samples that exhibit a large amount of coherent noise and a thickness larger than the depth of focus of the objective lens. As a demonstration, we performed a three-dimensional measurement of a fine metal mask with a reconstructed sectional phase image and filtering with a reconstructed amplitude image.

  10. Unilateral Hemithorax Opacification on Chest Radiograph : Comparison of Diagnostic Accuracy of Chest Ultrasonography with CT

    Energy Technology Data Exchange (ETDEWEB)

    Namkung, Sook; Lee, Kyung Soo; Kwon, O Jung [Samsung Medical Center, Seoul (Korea, Republic of); Chung, Seung Eun [Sangkye Paek Hospital, Seoul (Korea, Republic of)

    1996-12-15

    To compare the diagnostic accuracy of chest ultrasonography (US) with computed tomography (CT) inpatients with opacification more than one third of unilateral hemithorax on chest radiograph (CXR). Chest US and CT scans from 41 consecutive patients with opacification more than one third of unilateral hemithorax on CXR were prospectively evaluated by two independent radiologists. Each radiologist recorded 1) the nature of pleural effusion (transudate vs. exudate), 2) presence or absence of pulmonary lesion, 3) the characteristic of pulmonary lesion (consolidation or atelectasis and tumor), and 4) presence of solid pleural tumor. The diagnostic accuracy of chest US was compared with CT scan in patients with pleural, pulmonary or other disease. In 32 patients with pleural effusion, differentiation between transudate and exudate was feasible in 27 (84%) patients with US and 26 (81%) patients with CT. In 32 patients with pulmonary and other pleural diseases, sensitivity, specificity and accuracy of US in lesion detection were 86%, 75% and 83% respectively when CT was regarded as a diagnostic gold standard. The diagnostic accuracy of chest US is comparable to CT in patients with hemithorax opacification on CXR

  11. Unilateral Hemithorax Opacification on Chest Radiograph : Comparison of Diagnostic Accuracy of Chest Ultrasonography with CT

    International Nuclear Information System (INIS)

    Namkung, Sook; Lee, Kyung Soo; Kwon, O Jung; Chung, Seung Eun

    1996-01-01

    To compare the diagnostic accuracy of chest ultrasonography (US) with computed tomography (CT) inpatients with opacification more than one third of unilateral hemithorax on chest radiograph (CXR). Chest US and CT scans from 41 consecutive patients with opacification more than one third of unilateral hemithorax on CXR were prospectively evaluated by two independent radiologists. Each radiologist recorded 1) the nature of pleural effusion (transudate vs. exudate), 2) presence or absence of pulmonary lesion, 3) the characteristic of pulmonary lesion (consolidation or atelectasis and tumor), and 4) presence of solid pleural tumor. The diagnostic accuracy of chest US was compared with CT scan in patients with pleural, pulmonary or other disease. In 32 patients with pleural effusion, differentiation between transudate and exudate was feasible in 27 (84%) patients with US and 26 (81%) patients with CT. In 32 patients with pulmonary and other pleural diseases, sensitivity, specificity and accuracy of US in lesion detection were 86%, 75% and 83% respectively when CT was regarded as a diagnostic gold standard. The diagnostic accuracy of chest US is comparable to CT in patients with hemithorax opacification on CXR

  12. Comparison of digital tomosynthesis and chest radiography for the detection of pulmonary nodules: systematic review and meta-analysis.

    Science.gov (United States)

    Kim, Jun H; Lee, Kyung H; Kim, Kyoung-Tae; Kim, Hyun J; Ahn, Hyeong S; Kim, Yeo J; Lee, Ha Y; Jeon, Yong S

    2016-12-01

    To compare the diagnostic accuracy of digital tomosynthesis (DTS) with that of chest radiography for the detection of pulmonary nodules by meta-analysis. A systematic literature search was performed to identify relevant original studies from 1 January 1 1976 to 31 August 31 2016. The quality of included studies was assessed by quality assessment of diagnostic accuracy studies-2. Per-patient data were used to calculate the sensitivity and specificity and per-lesion data were used to calculate the detection rate. Summary receiver-operating characteristic curves were drawn for pulmonary nodule detection. 16 studies met the inclusion criteria. 1017 patients on a per-patient basis and 2159 lesions on a per-lesion basis from 16 eligible studies were evaluated. The pooled patient-based sensitivity of DTS was 0.85 [95% confidence interval (CI) 0.83-0.88] and the specificity was 0.95 (0.93-0.96). The pooled sensitivity and specificity of chest radiography were 0.47 (0.44-0.51) and 0.37 (0.34-0.40), respectively. The per-lesion detection rate was 2.90 (95% CI 2.63-3.19). DTS has higher diagnostic accuracy than chest radiography for detection of pulmonary nodules. Chest radiography has low sensitivity but similar specificity, comparable with that of DTS. Advances in knowledge: DTS has higher diagnostic accuracy than chest radiography for the detection of pulmonary nodules.

  13. Lung PET scan

    Science.gov (United States)

    ... Chest PET scan; Lung positron emission tomography; PET - chest; PET - lung; PET - tumor imaging; ... Grainger & Allison's Diagnostic Radiology: A Textbook of Medical Imaging . 6th ed. Philadelphia, ...

  14. Evaluation of entrance skin air kerma in digital and conventional pediatric chest examinations performed in Parana-Brazil

    International Nuclear Information System (INIS)

    Oliveira, Ana L.R.; Schelin, Hugo R.; Lunelli, Neuri; Paschuk, Sergei; Tilly Jr, Joao G.; Rogacheski, Enio; Khoury, Helen J.; Ferreira, Jose R.L.

    2008-01-01

    Full text: In pediatric radiology it is important to evaluate patient exposure due to the high sensitivity of his/her tissue and his/her relatively longer life expectancy, which may increase the chance of radiogenic cancer development. Special attention is necessary when digital systems are used since a high dosage level for the patient may go unnoticed because the resulting image is not overexposed. The purpose of this work is to evaluate the Entrance Skin Air Kerma (ESAK) in pediatric patients submitted to digital and conventional chest examinations performed in the State of Parana, located in the Southern region of Brazil. The survey was performed in two hospitals, one located in Curitiba, the Capital of the State (A) and the other in the city of Cascavel (B). Hospital A uses a conventional image system and Hospital B uses a CR digital system. The study group consisted of 50 patients (from 2 days to six years of age) undergoing chest PA/AP examinations. At the time of the examination, the exposure parameters (kV, mAs, focal-film distance, etc) and the gender, height, weight and age of the patient were recorded. To evaluate the Entrance Skin Air Kerma (ESAK), two TLD-100 dosimeters were packaged in a polyethylene case, heat sealed and placed on the surface of each patient. The TLDs were calibrated the Metrology Laboratory of Ionizing Radiation of the Federal University of Pernambuco. The TLDs were processed on a Victoreen readout system model 2800M. The results showed that the kilovoltage values used in Hospital A are in the range of 46 to 65 kV with the average value of 51,7 kV, while the ones in Hospital B are in the range of 65 to 80 kV. On the other hand, even though hospital B uses the CR system, the mAs used for the chest exams are 4 times higher than the ones used in Hospital A. For Hospital A, the mean ESAK value was 0.044 mGy (min: 0.08 and max 0.146 mGy), and for Hospital B the mean ESAK value was 0.368 mGy (min: 0.021 and max 1.21mGy). The results made

  15. MRI of the Chest

    Medline Plus

    Full Text Available ... a CD or uploaded to a digital cloud server. MRI of the chest gives detailed pictures of ... understanding of the possible charges you will incur. Web page review process: This Web page is reviewed ...

  16. MRI of the Chest

    Medline Plus

    Full Text Available ... to a CD or uploaded to a digital cloud server. MRI of the chest gives detailed pictures ... over time. top of page What are the benefits vs. risks? Benefits MRI is a noninvasive imaging ...

  17. MRI of the Chest

    Medline Plus

    Full Text Available ... to a CD or uploaded to a digital cloud server. MRI of the chest gives detailed pictures ... community, you can search the ACR-accredited facilities database . This website does not provide cost information. The ...

  18. MRI of the Chest

    Medline Plus

    Full Text Available ... copied to a CD or uploaded to a digital cloud server. MRI of the chest gives detailed ... sedative prior to your scheduled examination. Infants and young children usually require sedation or anesthesia to complete ...

  19. Advances in chest drain management in thoracic disease

    Science.gov (United States)

    George, Robert S.

    2016-01-01

    An adequate chest drainage system aims to drain fluid and air and restore the negative pleural pressure facilitating lung expansion. In thoracic surgery the post-operative use of the conventional underwater seal chest drainage system fulfills these requirements, however they allow great variability amongst practices. In addition they do not offer accurate data and they are often inconvenient to both patients and hospital staff. This article aims to simplify the myths surrounding the management of chest drains following chest surgery, review current experience and explore the advantages of modern digital chest drain systems and address their disease-specific use. PMID:26941971

  20. Quantitative assessment of the influence of anatomic noise on the detection of subtle lung nodule in digital chest radiography using fractal-feature distance

    International Nuclear Information System (INIS)

    Imai, Kuniharu; Ikeda, Mitsuru; Enchi, Yukihiro; Niimi, Takanaga

    2008-01-01

    Purpose: To confirm whether or not the influence of anatomic noise on the detection of nodules in digital chest radiography can be evaluated by the fractal-feature distance. Materials and methods: We used the square images with and without a simulated nodule which were generated in our previous observer performance study; the simulated nodule was located on the upper margin of a rib, the inside of a rib, the lower margin of a rib, or the central region between two adjoining ribs. For the square chest images, fractal analysis was conducted using the virtual volume method. The fractal-feature distances between the considered and the reference images were calculated using the pseudo-fractal dimension and complexity, and the square images without the simulated nodule were employed as the reference images. We compared the fractal-feature distances with the observer's confidence level regarding the presence of a nodule in plain chest radiograph. Results: For all square chest images, the relationships between the length of the square boxes and the mean of the virtual volumes were linear on a log-log scale. For all types of the simulated nodules, the fractal-feature distance was the highest for the simulated nodules located on the central region between two adjoining ribs and was the lowest for those located in the inside of a rib. The fractal-feature distance showed a linear relation to an observer's confidence level. Conclusion: The fractal-feature distance would be useful for evaluating the influence of anatomic noise on the detection of nodules in digital chest radiography

  1. Multi-Institutional Evaluation of Digital Tomosynthesis, Dual-Energy Radiography, and Conventional Chest Radiography for the Detection and Management of Pulmonary Nodules.

    Science.gov (United States)

    Dobbins, James T; McAdams, H Page; Sabol, John M; Chakraborty, Dev P; Kazerooni, Ella A; Reddy, Gautham P; Vikgren, Jenny; Båth, Magnus

    2017-01-01

    Purpose To conduct a multi-institutional, multireader study to compare the performance of digital tomosynthesis, dual-energy (DE) imaging, and conventional chest radiography for pulmonary nodule detection and management. Materials and Methods In this binational, institutional review board-approved, HIPAA-compliant prospective study, 158 subjects (43 subjects with normal findings) were enrolled at four institutions. Informed consent was obtained prior to enrollment. Subjects underwent chest computed tomography (CT) and imaging with conventional chest radiography (posteroanterior and lateral), DE imaging, and tomosynthesis with a flat-panel imaging device. Three experienced thoracic radiologists identified true locations of nodules (n = 516, 3-20-mm diameters) with CT and recommended case management by using Fleischner Society guidelines. Five other radiologists marked nodules and indicated case management by using images from conventional chest radiography, conventional chest radiography plus DE imaging, tomosynthesis, and tomosynthesis plus DE imaging. Sensitivity, specificity, and overall accuracy were measured by using the free-response receiver operating characteristic method and the receiver operating characteristic method for nodule detection and case management, respectively. Results were further analyzed according to nodule diameter categories (3-4 mm, >4 mm to 6 mm, >6 mm to 8 mm, and >8 mm to 20 mm). Results Maximum lesion localization fraction was higher for tomosynthesis than for conventional chest radiography in all nodule size categories (3.55-fold for all nodules, P chest radiography for all nodules (1.49-fold, P chest radiography, as given by the area under the receiver operating characteristic curve (1.23-fold, P chest radiography or tomosynthesis. Conclusion Tomosynthesis outperformed conventional chest radiography for lung nodule detection and determination of case management; DE imaging did not show significant differences over conventional chest

  2. Reliability of capturing foot parameters using digital scanning and the neutral suspension casting technique

    Science.gov (United States)

    2011-01-01

    Background A clinical study was conducted to determine the intra and inter-rater reliability of digital scanning and the neutral suspension casting technique to measure six foot parameters. The neutral suspension casting technique is a commonly utilised method for obtaining a negative impression of the foot prior to orthotic fabrication. Digital scanning offers an alternative to the traditional plaster of Paris techniques. Methods Twenty one healthy participants volunteered to take part in the study. Six casts and six digital scans were obtained from each participant by two raters of differing clinical experience. The foot parameters chosen for investigation were cast length (mm), forefoot width (mm), rearfoot width (mm), medial arch height (mm), lateral arch height (mm) and forefoot to rearfoot alignment (degrees). Intraclass correlation coefficients (ICC) with 95% confidence intervals (CI) were calculated to determine the intra and inter-rater reliability. Measurement error was assessed through the calculation of the standard error of the measurement (SEM) and smallest real difference (SRD). Results ICC values for all foot parameters using digital scanning ranged between 0.81-0.99 for both intra and inter-rater reliability. For neutral suspension casting technique inter-rater reliability values ranged from 0.57-0.99 and intra-rater reliability values ranging from 0.36-0.99 for rater 1 and 0.49-0.99 for rater 2. Conclusions The findings of this study indicate that digital scanning is a reliable technique, irrespective of clinical experience, with reduced measurement variability in all foot parameters investigated when compared to neutral suspension casting. PMID:21375757

  3. Impact of digital intraoral scan strategies on the impression accuracy using the TRIOS Pod scanner.

    Science.gov (United States)

    Müller, Philipp; Ender, Andreas; Joda, Tim; Katsoulis, Joannis

    2016-04-01

    Little information is available on the impact of different scan strategies on the accuracy of full-arch scans with intraoral scanners. The aim of this in-vitro study was to investigate the trueness and precision of full-arch maxillary digital impressions comparing three scan strategies. Three scan strategies (A, B, and C) were applied each five times on one single model (A, first buccal surfaces, return from occlusal-palatal; B, first occlusal-palatal, return buccal; C, S-type one-way). The TRIOS Pod scanner (3shape, Copenhagen, Denmark) with a color detector was used for these digital impressions. A cast of a maxillary dentate jaw was fabricated and scanned with an industrial reference scanner. This full-arch data record was digitally superimposed with the test scans (trueness) and within-group comparison was performed for each group (precision). The values within the 90/10 percentiles from the digital superimposition were used for calculation and group comparisons with nonparametric tests (ANOVA, post-hoc Bonferroni). The trueness (mean ± standard deviation) was 17.9 ± 16.4 μm for scan strategy A, 17.1 ± 13.7 μm for B, and 26.8 ± 14.7 μm for C without statistically significant difference. The precision was lowest for scan strategy A (35.0 ± 51.1 μm) and significantly different to B (7.9 ± 5.6 μm) and C (8.5 ± 6.3 μm). Scan strategy B may be recommended as it provides the highest trueness and precision in full-arch scans and therefore minimizes inaccuracies in the final reconstruction.

  4. A digital gigapixel large-format tile-scan camera.

    Science.gov (United States)

    Ben-Ezra, M

    2011-01-01

    Although the resolution of single-lens reflex (SLR) and medium-format digital cameras has increased in recent years, applications for cultural-heritage preservation and computational photography require even higher resolutions. Addressing this issue, a large-format cameras' large image planes can achieve very high resolution without compromising pixel size and thus can provide high-quality, high-resolution images.This digital large-format tile scan camera can acquire high-quality, high-resolution images of static scenes. It employs unique calibration techniques and a simple algorithm for focal-stack processing of very large images with significant magnification variations. The camera automatically collects overlapping focal stacks and processes them into a high-resolution, extended-depth-of-field image.

  5. Full arch scans: conventional versus digital impressions--an in-vitro study.

    Science.gov (United States)

    Ender, A; Mehl, A

    2011-01-01

    The digital intraoral impression has become a central part of the CAD/CAM technique. The objective of the present study was to compare the accuracy (trueness and precision) of digital impressions of the full arch with that of conventional impressions on the in-vitro model. For this purpose, a master model was acquired with a new reference scanning process, the measuring trueness of which was +/- 4.1 microm and the precision +/- 2.5 microm. On the one hand, conventional impressions and then plaster models (n = 5) were produced from this master model, and on the other hand, digital impressions were made with the Cerec AC Bluecam and the Lava COS system (each n = 5). The plaster models were also scanned with the reference scanner. The available data records were superimposed and the differences determined. The deviation from the master model defines the trueness of the impression method. The deviations of the models among one another demonstrate the precision of the method. The trueness of the impressions was 55 +/- 21.8 microm in the conventional impression group, for digital impressions with Cerec Bluecam it was 49 +/- 14.2 microm and for digital impressions with Lava COS 40.3 +/- 14.1 microm. The precision was 61.3 +/- 17.9 microm for conventional impression with Impregum, 30.9 +/- 7.1 microm for digital impression with the Cerec Bluecam and 60.1 +/- 31.3 microm for digital impression with Lava COS. These in-vitro results show that accuracy of the digital impression is similar to that of the conventional impression. These results will have to be confirmed in further clinical studies.

  6. A simplified digital lock-in amplifier for the scanning grating spectrometer.

    Science.gov (United States)

    Wang, Jingru; Wang, Zhihong; Ji, Xufei; Liu, Jie; Liu, Guangda

    2017-02-01

    For the common measurement and control system of a scanning grating spectrometer, the use of an analog lock-in amplifier requires complex circuitry and sophisticated debugging, whereas the use of a digital lock-in amplifier places a high demand on the calculation capability and storage space. In this paper, a simplified digital lock-in amplifier based on averaging the absolute values within a complete period is presented and applied to a scanning grating spectrometer. The simplified digital lock-in amplifier was implemented on a low-cost microcontroller without multipliers, and got rid of the reference signal and specific configuration of the sampling frequency. Two positive zero-crossing detections were used to lock the phase of the measured signal. However, measurement method errors were introduced by the following factors: frequency fluctuation, sampling interval, and integer restriction of the sampling number. The theoretical calculation and experimental results of the signal-to-noise ratio of the proposed measurement method were 2055 and 2403, respectively.

  7. Automatic scanning of Cerenkov light photograms from a multistep avalanche chamber using a television digitizer

    International Nuclear Information System (INIS)

    Vascon, M.; Zanella, G.

    1980-01-01

    A television digitizer and its application to automatic scanning of Cerenkov imaging using the multistep avalanche chamber in front of optical spark chamber are described. The results are of interest in the adoption of the automatic scanning of photographic plates of these events or for the on-line application of the television digitizer itself. (orig.)

  8. Chest pain

    International Nuclear Information System (INIS)

    Martinez A, Juan Carlos; Saenz M, Oscar; Martinez M, Camilo; Gonzales A Francisco; Nicolas R, Jose; Vergara V, Erika P; Pereira G, Alberto M

    2010-01-01

    In emergency departments, chest pain is one of the leading motives of consultation. We thus consider it important to review aspects such as its classification, causes, and clinical profiles. Initial assessment should include a full clinical history comprising thorough anamnesis and physical examination. Adequate interpretation of auxiliary tests, ordered in accordance with suspected clinical conditions, should lead to accurate diagnosis. We highlight certain symptoms and clinical signs, ECG and X-ray findings, cardiac bio markers, arterial blood gases, and CT-scanning. Scores of severity and prognosis such as TIMI are assessed. Optimal treatment of the clinical conditions leading to chest pain depends on adequate initial approach and assessment.

  9. Digital chest tomosynthesis: the 2017 updated review of an emerging application

    Science.gov (United States)

    Ferrari, Arianna; Bertolaccini, Luca; Solli, Piergiorgio; Di Salvia, Paola Oriana

    2018-01-01

    Lung cancer is the leading cause of cancer death and second most common cancer among both men and women, but most of them are detected when patients become symptomatic and in late-stage. Chest radiography (CR) is a basic technique for the investigation of lung cancer and has the benefit of convenience and low radiation dose, but detection of malignancy is often difficult. The introduction of computed tomography (CT) for screening has increased the proportion of lung cancer detected but with higher exposure dose and higher costs. Digital chest tomosynthesis (DCT), a tomographic technique, may offer an alternative to CT. DCT uses a conventional radiograph tube, a flat-panel detector, a computer-controlled tube mover and reconstruction algorithms to produce section images. It shows promise in the detection of potentially malignant lung nodules, with higher sensibility than CR, and is emerging as a low-dose and low-cost alternative to CT to improve treatment decisions. In fact, an increasing number of researchers are showing that tomosynthesis could have a role in the detection of lung cancer, in addition to its present role in breast screening. However, DCT offers some limitations, such as limited depth resolution, which may explain the difficulty in detecting pathologies in the subpleural region and the occurrence of artefacts from medical devices. Once solved these limitations and once more studies supporting its use will be available, DCT could become the first-line lung cancer screening tool among patients at considerable risk of lung cancer. PMID:29666814

  10. Implementation of a dedicated digital projectional radiographic system in thoracic imaging

    International Nuclear Information System (INIS)

    Aberle, D.R.; Batra, P.; Hayrapetian, A.S.; Brown, K.; Morioka, C.A.; Steckel, R.J.

    1988-01-01

    An integrated digital radiographic system was evaluated with respect to image quality and impact on diagnosis relative to conventional chest radiographs for a variety of focal and diffuse lung processes. Digital images were acquired with a stimulable phosphor plate detector that was scanned by a semiconductor laser for immediate digitalization to a 2,048 X 2,464 X 10-bit image. Digital images were displayed on a 2,048-line monitor and printed on 14 X 17-inch film with use of a laser film printer (Kodak). Preliminary results with this system, including the effects of user interaction with the display monitor, inverse intensity display, and regional magnification techniques, indicate that it may be successfully implemented for thoracic imaging

  11. Accuracy and Reliability of a Novel Method for Fusion of Digital Dental Casts and Cone Beam Computed Tomography Scans

    Science.gov (United States)

    Rangel, Frits A.; Maal, Thomas J. J.; Bronkhorst, Ewald M.; Breuning, K. Hero; Schols, Jan G. J. H.; Bergé, Stefaan J.; Kuijpers-Jagtman, Anne Marie

    2013-01-01

    Several methods have been proposed to integrate digital models into Cone Beam Computed Tomography scans. Since all these methods have some drawbacks such as radiation exposure, soft tissue deformation and time-consuming digital handling processes, we propose a new method to integrate digital dental casts into Cone Beam Computed Tomography scans. Plaster casts of 10 patients were randomly selected and 5 titanium markers were glued to the upper and lower plaster cast. The plaster models were scanned, impressions were taken from the plaster models and the impressions were also scanned. Linear measurements were performed on all three models, to assess accuracy and reproducibility. Besides that, matching of the scanned plaster models and scanned impressions was done, to assess the accuracy of the matching procedure. Results show that all measurement errors are smaller than 0.2 mm, and that 81% is smaller than 0.1 mm. Matching of the scanned plaster casts and scanned impressions show a mean error between the two surfaces of the upper arch of 0.14 mm and for the lower arch of 0.18 mm. The time needed for reconstructing the CBCT scans to a digital patient, where the impressions are integrated into the CBCT scan of the patient takes about 15 minutes, with little variance between patients. In conclusion, we can state that this new method is a reliable method to integrate digital dental casts into CBCT scans. As far as radiation exposure, soft tissue deformation and digital handling processes are concerned, it is a significant improvement compared to the previously published methods. PMID:23527111

  12. Accuracy and reliability of a novel method for fusion of digital dental casts and Cone Beam Computed Tomography scans.

    Directory of Open Access Journals (Sweden)

    Frits A Rangel

    Full Text Available Several methods have been proposed to integrate digital models into Cone Beam Computed Tomography scans. Since all these methods have some drawbacks such as radiation exposure, soft tissue deformation and time-consuming digital handling processes, we propose a new method to integrate digital dental casts into Cone Beam Computed Tomography scans. Plaster casts of 10 patients were randomly selected and 5 titanium markers were glued to the upper and lower plaster cast. The plaster models were scanned, impressions were taken from the plaster models and the impressions were also scanned. Linear measurements were performed on all three models, to assess accuracy and reproducibility. Besides that, matching of the scanned plaster models and scanned impressions was done, to assess the accuracy of the matching procedure. Results show that all measurement errors are smaller than 0.2 mm, and that 81% is smaller than 0.1 mm. Matching of the scanned plaster casts and scanned impressions show a mean error between the two surfaces of the upper arch of 0.14 mm and for the lower arch of 0.18 mm. The time needed for reconstructing the CBCT scans to a digital patient, where the impressions are integrated into the CBCT scan of the patient takes about 15 minutes, with little variance between patients. In conclusion, we can state that this new method is a reliable method to integrate digital dental casts into CBCT scans. As far as radiation exposure, soft tissue deformation and digital handling processes are concerned, it is a significant improvement compared to the previously published methods.

  13. Quantitative evaluation of anatomical noise in chest digital tomosynthesis, digital radiography, and computed tomography

    International Nuclear Information System (INIS)

    Lee, D.; Kim, D.; Choi, S.; Kim, H.-J.; Choi, S.; Lee, H.

    2017-01-01

    Lung cancer is currently the worldwide leading cause of death from cancer. Thus, detection of lung cancer at its early stages is critical for improving the survival rate of patients. Chest digital tomosynthesis (CDT) is a recently developed imaging modality, combining many advantages of digital radiography (DR) and computed tomography (CT). This method has the potential to be widely used in the clinical setting. In this study, we introduce a developed CDT R/F system and compare its image quality with those of DR and CT, especially with respect to anatomical noise and lung nodule conspicuity, for LUNGMAN phantoms. The developed CDT R/F system consists of a CsI scintillator flat panel detector, X-ray tube, and tomosynthesis data acquisition geometry. For CDT R/F imaging, 41 projections were acquired at different angles, over the ± 20° angular range, in a linear translation geometry. To evaluate the clinical effectiveness of the CDT R/F system, the acquired images were compared with CT (Philips brilliance CT 64, Philips healthcare, U.S.) and DR (ADR-M, LISTEM, Korea) phantom images in terms of the anatomical noise power spectrum (aNPS). DR images exhibited low conspicuity for a small-size lung nodule, while CDT R/F and CT exhibited relatively high sensitivity for all lung nodule sizes. The aNPS of the CDT R/F system was better than that of DR, by resolving anatomical overlapping problems. In conclusion, the developed CDT R/F system is likely to contribute to early diagnosis of lung cancer, while requiring a relatively low patient dose, compared with CT.

  14. Quantitative evaluation of anatomical noise in chest digital tomosynthesis, digital radiography, and computed tomography

    Science.gov (United States)

    Lee, D.; Choi, S.; Lee, H.; Kim, D.; Choi, S.; Kim, H.-J.

    2017-04-01

    Lung cancer is currently the worldwide leading cause of death from cancer. Thus, detection of lung cancer at its early stages is critical for improving the survival rate of patients. Chest digital tomosynthesis (CDT) is a recently developed imaging modality, combining many advantages of digital radiography (DR) and computed tomography (CT). This method has the potential to be widely used in the clinical setting. In this study, we introduce a developed CDT R/F system and compare its image quality with those of DR and CT, especially with respect to anatomical noise and lung nodule conspicuity, for LUNGMAN phantoms. The developed CDT R/F system consists of a CsI scintillator flat panel detector, X-ray tube, and tomosynthesis data acquisition geometry. For CDT R/F imaging, 41 projections were acquired at different angles, over the ± 20° angular range, in a linear translation geometry. To evaluate the clinical effectiveness of the CDT R/F system, the acquired images were compared with CT (Philips brilliance CT 64, Philips healthcare, U.S.) and DR (ADR-M, LISTEM, Korea) phantom images in terms of the anatomical noise power spectrum (aNPS). DR images exhibited low conspicuity for a small-size lung nodule, while CDT R/F and CT exhibited relatively high sensitivity for all lung nodule sizes. The aNPS of the CDT R/F system was better than that of DR, by resolving anatomical overlapping problems. In conclusion, the developed CDT R/F system is likely to contribute to early diagnosis of lung cancer, while requiring a relatively low patient dose, compared with CT.

  15. The effect of Tc99m Sestamibi scans during acute chest pain on clinical management

    International Nuclear Information System (INIS)

    Baldey, A.; Cameron, P.; Grigg, L.; Knott, J.; Better, N.

    1998-01-01

    Full text: The aim of this study is to assess whether the increased sensitivity and specificity of Tc99m sestamibi scans, during acute chest pain, will lead to alteration in clinical management and potential cost saving in an Australian population. Consecutive patients who presented with acute chest pain were injected 800 MBq of Tc99m sestamibi during pain (Hot MlBI) and SPECT imaging performed 1-6 hours later. The population was those only with a 'intermediate risk' of myocardial ischaemia The patients included in patients, those in the Emergency Department, and those with a previous history of cardiac disease. 25% of patients required a second, pain free study the following day to differentiate acute ischaemia from prior infarction. A question sheet was filled out by the requesting physician prior to the study indicating the likelihood of cardiac disease and the proposed management if no 'Hot MIBI' scan was available. The treatment that the patient subsequently received was ascertained from the patient's medical record. Of the 28 patients, a prediction whether to or not to proceed to coronary angiography was made in 13 patients prior to the MIBI study being performed. Of the 13, 5 would have had coronary angiography performed. and in all 5, the decision to proceed to coronary angiography was averted by the 'Hot MIBI'. Of note, 3 patients were admitted purely because of an abnormal 'Hot MIBI'. The 'Hot MIBI' was able to reduce coronary care admissions by 83% reduce all admissions by 17%, and avert coronary angiography in 38% of patients. In this intermediate risk category patient, this translates to not only admissions saved but potential cost saving

  16. [Conventional radiology, digital radiology with photostimulable phosphor, laser digitalization of thoracic radiographic films at the bedside. A comparative study].

    Science.gov (United States)

    Miceli, M; Stamati, R; Burci, P; Guidarelli, G; Sartoni Galloni, S

    1992-10-01

    The bedside chest images obtained with conventional radiology and with "on line" and "off line" digital modalities were compared to evaluate the respective capabilities in visualizing chest anatomical structures. Seventy patients were submitted to bedside chest examinations with a portable unit; both a conventional film and a digital system (PCR Graphics 1, Philips) with photostimulable phosphor imaging plate were fitted in the radiographic cassette. The former was digitized using an "off line" laser beam unit (FD 2000, Dupont); the latter was subsequently postprocessed by modifying contrast, optical density and spatial frequencies. Thus, 4 different viewing modalities were obtained for each examination: a) conventional radiography; b) standard digital radiography; c) postprocessed digital radiography; d) digitized conventional radiography. Detectability rates of chest anatomical structures were analyzed by 4 independent radiologists on the different images and expressed by a score 1-4. The values were always higher with digital modalities than with the conventional one and the differences were statistically significant (Student's t-test modified by Bonferroni). In particular, the greatest difference was found between c) and a) in retrocardiac lung parenchyma and in skeletal structures, in favour of c). Concerning the comparative adequacy of the various digital modalities, higher detectability rates of chest anatomical structures were obtained with c), but also with b), than with d).

  17. Modelling the detachment dependence on strike point location in the small angle slot divertor (SAS) with SOLPS

    Science.gov (United States)

    Casali, Livia; Covele, Brent; Guo, Houyang

    2017-10-01

    The new Small Angle Slot (SAS) divertor in DIII-D is characterized by a shallow-angle target enclosed by a slot structure about the strike point (SP). SOLPS modelling results of SAS have demonstrated divertor closure's utility in widening the range of acceptable densities for adequate heat handling. An extensive database of runs has been built to study the detachment dependence on SP location in SAS. Density scans show that lower Te at lower upstream density occur when the SP is at the critical location in the slot. The cooling front spreads across the entire target at higher densities, in agreement with experimental Langmuir probe measurements. A localized increase of the atomic and molecular density takes place near the SP, which reduces the target incident power density and facilitates detachment at lower upstream density. Systematic scans of variables such as power, transport, and viscosity have been carried out to assess the detachment sensitivity. Therein, a positive role of the viscosity is found. This work supported by DOE Contract Number DE-FC02-04ER54698.

  18. Digital examination and transvaginal scan - competing or complementary for predicting preterm birth?

    DEFF Research Database (Denmark)

    Reiter, Eva; Nielsen, Kurt Aagaard; Fedder, Jens

    2012-01-01

    The transvaginal ultrasonographic cervix scan has partly replaced digital examination for diagnosing preterm birth; conflicting results are reported about their respective contribution to birth prediction....

  19. Digital imaging in diagnostic radiology

    International Nuclear Information System (INIS)

    Newell, J.D. Jr.; Kelsey, C.A.

    1990-01-01

    This monograph on digital imaging provides a basic overview of this field at the present time. This paper covers clinical application, including subtraction angiography; chest radiology; genitourinary, gastrointestinal, and breast radiology; and teleradiology. The chest section also includes an explanation of multiple beam equalization radiography. The remaining chapters discuss some of the technical aspects of digital radiology. It includes the basic technology of digital radiography, image compression, and reconstruction information on the economics of digital radiography

  20. Myocardial contusion following nonfatal blunt chest trauma

    International Nuclear Information System (INIS)

    Kumar, S.A.; Puri, V.K.; Mittal, V.K.; Cortez, J.

    1983-01-01

    Currently available diagnostic techniques for myocardial contusion following blunt chest trauma were evaluated. We investigated 30 patients prospectively over a period of 1 year for the presence of myocardial contusion. Among the 30 patients, eight were found to have myocardial contusion on the basis of abnormal electrocardiograms, elevated creatine phosphokinase MB fraction (CPK-MB), and positive myocardial scan. Myocardial scan was positive in seven of eight patients (87.5%). CPK-MB fraction was elevated in four of eight patients (50%). Definitive electrocardiographic changes were seen in only two of eight patients (25%). It appears that myocardial scan using technetium pyrophosphate and CPK-MB fraction determinations are the most reliable aids in diagnosis of myocardial contusion following blunt chest trauma

  1. Slot-coupled CW standing wave accelerating cavity

    Energy Technology Data Exchange (ETDEWEB)

    Wang, Shaoheng; Rimmer, Robert; Wang, Haipeng

    2017-05-16

    A slot-coupled CW standing wave multi-cell accelerating cavity. To achieve high efficiency graded beta acceleration, each cell in the multi-cell cavity may include different cell lengths. Alternatively, to achieve high efficiency with acceleration for particles with beta equal to 1, each cell in the multi-cell cavity may include the same cell design. Coupling between the cells is achieved with a plurality of axially aligned kidney-shaped slots on the wall between cells. The slot-coupling method makes the design very compact. The shape of the cell, including the slots and the cone, are optimized to maximize the power efficiency and minimize the peak power density on the surface. The slots are non-resonant, thereby enabling shorter slots and less power loss.

  2. Intra-abdominal fat area measurement using chest CT data

    International Nuclear Information System (INIS)

    Moriya, Hiroshi; Midorikawa, Shigeo; Hashimoto, Kouji; Ishii, Akira; Saitou, Kumi; Andou, Tomonori; Kitamura, Naoko; Sakuma, Koutarou

    2007-01-01

    Intra-abdominal fat obesity, which is linked with the metabolic syndrome, is usually characterized by measuring intra-abdominal fat area at the umbilical level of abdominal CT scan. In recent year, the chances of chest CT scanning are increased, as lung cancer screening survey or individual medical examination. Thus, we presented a method of measuring the areas of intra-abdominal fat and subcutaneous fat at the lower slice of chest CT scan. Fat areas found with this method were significantly correlated with those obtained at the umbilical level. (author)

  3. Coupling slots without shunt impedance drop

    International Nuclear Information System (INIS)

    Balleyguier, P.

    1996-01-01

    It is well known that coupling slots between adjacent cells in a π-mode structure reduce shunt impedance per unit length with respect to single cell cavities. To design optimized coupling slots, one has to answer the following question: for a given coupling factor, what shape, dimension, position and number of slots lead to the lowest shunt impedance drop? A numerical study using the 3D code MAFIA has been carried out. The aim was to design the 352 MHz cavities for the high intensity proton accelerator of the TRISPAL project. The result is an unexpected set of four 'petal' slots. Such slots should lead to a quasi-negligible drop in shunt impedance: about -1% on average, for particle velocity from 0.4 c to 0.8 c. (author)

  4. CT findings of chest trauma

    International Nuclear Information System (INIS)

    Kim, Young Tong; Kim Young Il

    1998-01-01

    Trauma is the third leading cause of death, irrespective of age, and the leading cause of death in persons under 40 persons under 40 years of age. Most pleural, pulmonary, mediastinal, and diaphragmatic injuries are not seen on conventional chest radiographs, or are underestimated. In patients with chest trauma, CT scanning is an effective and sensitive method of detecting thoracic injuries and provides accurate information regarding their pattern and extent. (author). 5 refs., 17 figs

  5. Effect of display type and room illuminance in chest radiographs

    International Nuclear Information System (INIS)

    Liukkonen, Esa; Jartti, Airi; Haapea, Marianne; Oikarinen, Heljae; Ahvenjaervi, Lauri; Mattila, Seija; Nevala, Terhi; Palosaari, Kari; Perhomaa, Marja; Nieminen, Miika T.

    2016-01-01

    To compare diagnostic accuracy in the detection of subtle chest lesions on digital chest radiographs using medical-grade displays, consumer-grade displays, and tablet devices under bright and dim ambient light. Five experienced radiologists independently assessed 50 chest radiographs (32 with subtle pulmonary findings and 18 without apparent findings) under bright (510 lx) and dim (16 lx) ambient lighting. Computed tomography was used as the reference standard for interstitial and nodular lesions and follow-up chest radiograph for pneumothorax. Diagnostic accuracy and sensitivity were calculated for assessments carried out in all displays and compared using the McNemar test. The level of significance was set to p < 0.05. Significant differences in sensitivity between the assessments under bright and dim lighting were found among consumer-grade displays in interstitial opacities with, and in pneumothorax without, Digital Imaging and Communication in Medicine-Grayscale Standard Display Function (DICOM-GSDF) calibration. Compared to 6 megapixel (MP) display under bright lighting, sensitivity in pneumothorax was lower in the tablet device and the consumer-grade display. Sensitivity in interstitial opacities was lower in the DICOM-GSDF calibrated consumer-grade display. A consumer-grade display with or without DICOM-GSDF calibration or a tablet device is not suitable for reading digital chest radiographs in bright lighting. No significant differences were observed between five displays in dim light. (orig.)

  6. Multiple-energy tissue-cancellation applications of a digital beam attenuator to chest radiography

    International Nuclear Information System (INIS)

    Dobbins, J.T. III.

    1985-01-01

    The digitally-formed primary beam attenuator (DBA) spatially modulates the x-ray fluence incident upon the patient to selectively attenuate regions of interest. The DBA attenuating mask is constructed from CeO 2 powder by a modified printing technique and uses image information from an initial low-dose exposure. Two tissue-cancellation imaging techniques are investigated with the DBA: (1) energy-dependent information is used to form a beam attenuator that attenuates specific tissues in the primary x-ray beam for tissue-cancelled film radiography; (2) the beam attenuator is used to improve image signal-to-noise and scattered radiation properties in traditional energy-subtraction tissue-cancellation imaging with digital detectors. The tissue-cancellation techniques in the primary x-ray beam were capable of adequately removing either soft-tissue or bone from the final compensated film radiograph when using a phantom with well defined soft-tissue and bone sections. However, when tried on an anthropomorphic chest phantom the results were adequate for cancellation of large soft tissue structures, but unsatisfactory for cancellation of bony structures such as the ribs, because of the limited spatial frequency content of the attenuating mask. The second technique (with digital detectors) showed improved uniformity of image signal-to-noise and a two-fold increase in soft-tissue nodule contrast due to improved scattered radiation properties. The tissue-cancelled images contained residual image contributions from the presence of the attenuating mask, but this residual may be correctable by future algorithms

  7. Slot-Coupled Barbel Antenna

    DEFF Research Database (Denmark)

    Jørgensen, Kasper Lüthje; Jakobsen, Kaj Bjarne

    2016-01-01

    A novel slot-coupled barbel antenna is designed and analyzed. A sensitivity analysis performed in order to improve the bandwidth, while the center frequency is kept constant.......A novel slot-coupled barbel antenna is designed and analyzed. A sensitivity analysis performed in order to improve the bandwidth, while the center frequency is kept constant....

  8. G-Tunnel pressurized slot-testing preparations

    International Nuclear Information System (INIS)

    Zimmerman, R.M.; Sifre-Soto, C.; Mann, K.L.; Bellman, R.A. Jr.; Luker, S.; Dodds, D.J.

    1992-04-01

    Designers and analysts of radioactive waste repositories must be able to predict the mechanical behavior of the host rock. Sandia National laboratories elected to conduct a development program on pressurized slot testing and featured (1) development of an improved method to cut slots using a chain saw with diamond-tipped cutters, (2) measurements useful for determining in situ stresses normal to slots, (3) measurements applicable for determining the in situ modulus of deformation parallel to a drift surface, and (4) evaluations of the potentials of pressurized slot strength testing. This report describes the preparations leading to the measurements and evaluations

  9. Optical Slot-Waveguide Based Biochemical Sensors

    Directory of Open Access Journals (Sweden)

    Carlos Angulo Barrios

    2009-06-01

    Full Text Available Slot-waveguides allow light to be guided and strongly confined inside a nanometer-scale region of low refractive index. Thus stronger light-analyte interaction can be obtained as compared to that achievable by a conventional waveguide, in which the propagating beam is confined to the high-refractive-index core of the waveguide. In addition, slot-waveguides can be fabricated by employing CMOS compatible materials and technology, enabling miniaturization, integration with electronic, photonic and fluidic components in a chip, and mass production. These advantages have made the use of slot-waveguides for highly sensitive biochemical optical integrated sensors an emerging field. In this paper, recent achievements in slot-waveguide based biochemical sensing will be reviewed. These include slot-waveguide ring resonator based refractometric label-free biosensors, label-based optical sensing, and nano-opto-mechanical sensors.

  10. Usefulness of computerized method for lung nodule detection on digital chest radiographs using similar subtraction images from different patients

    International Nuclear Information System (INIS)

    Aoki, Takatoshi; Oda, Nobuhiro; Yamashita, Yoshiko; Yamamoto, Keiji; Korogi, Yukunori

    2012-01-01

    Purpose: The purpose of this study is to evaluate the usefulness of a novel computerized method to select automatically the similar chest radiograph for image subtraction in the patients who have no previous chest radiographs and to assist the radiologists’ interpretation by presenting the “similar subtraction image” from different patients. Materials and methods: Institutional review board approval was obtained, and the requirement for informed patient consent was waived. A large database of approximately 15,000 normal chest radiographs was used for searching similar images of different patients. One hundred images of candidates were selected according to two clinical parameters and similarity of the lung field in the target image. We used the correlation value of chest region in the 100 images for searching the most similar image. The similar subtraction images were obtained by subtracting the similar image selected from the target image. Thirty cases with lung nodules and 30 cases without lung nodules were used for an observer performance test. Four attending radiologists and four radiology residents participated in this observer performance test. Results: The AUC for all radiologists increased significantly from 0.925 to 0.974 with the CAD (P = .004). When the computer output images were available, the average AUC for the residents was more improved (0.960 vs. 0.890) than for the attending radiologists (0.987 vs. 0.960). Conclusion: The novel computerized method for lung nodule detection using similar subtraction images from different patients would be useful to detect lung nodules on digital chest radiographs, especially for less experienced readers.

  11. Dose and perceived image quality in chest radiography

    International Nuclear Information System (INIS)

    Veldkamp, Wouter J.H.; Kroft, Lucia J.M.; Geleijns, Jacob

    2009-01-01

    Chest radiography is the most commonly performed diagnostic X-ray examination. The radiation dose to the patient for this examination is relatively low but because of its frequent use, the contribution to the collective dose is considerable. Consequently, optimization of dose and image quality offers a challenging area of research. In this article studies on dose reduction, different detector technologies, optimization of image acquisition and new technical developments in image acquisition and post processing will be reviewed. Studies indicate that dose reduction in PA chest images to at least 50% of commonly applied dose levels does not affect diagnosis in the lung fields; however, dose reduction in the mediastinum, upper abdomen and retrocardiac areas appears to directly deteriorate diagnosis. In addition to patient dose, also the design of the various digital detectors seems to have an effect on image quality. With respect to image acquisition, studies showed that using a lower tube voltage improves visibility of anatomical structures and lesions in digital chest radiographs but also increases the disturbing appearance of ribs. New techniques that are currently being evaluated are dual energy, tomosynthesis, temporal subtraction and rib suppression. These technologies may improve diagnostic chest X-ray further. They may for example reduce the negative influence of over projection of ribs, referred to as anatomic noise. In chest X-ray this type of noise may be the dominating factor in the detection of nodules. In conclusion, optimization and new developments will enlarge the value of chest X-ray as a mainstay in the diagnosis of chest diseases.

  12. Use of morphologic filters in the computerized detection of lung nodules in digital chest images

    International Nuclear Information System (INIS)

    Yoshimura, H.; Giger, M.L.; Doi, K.; Ahn, N.; MacMahon, H.

    1989-01-01

    The authors have previously described a computerized scheme for the detection of lung nodules based on a difference-image approach, which had a detection accuracy of 70% with 7--8 false positives per image. Currently, they are investigating morphologic filters for the further enhancement/suppression of nodule-signals and the removal of false-positives. Gray-level morphologic filtering is performed on clinical chest radiographs digitized with an optical drum scanner. Various shapes and sequences of erosion and dilation filters (i.e., determination of the minimum and maximum gray levels, respectively) were examined for signal enhancement and suppression for sue in the difference- image approach

  13. Diagnostic impact of digital tomosynthesis in oncologic patients with suspected pulmonary lesions on chest radiography.

    Science.gov (United States)

    Quaia, Emilio; Baratella, Elisa; Poillucci, Gabriele; Gennari, Antonio Giulio; Cova, Maria Assunta

    2016-08-01

    To assess the actual diagnostic impact of digital tomosynthesis (DTS) in oncologic patients with suspected pulmonary lesions on chest radiography (CXR). A total of 237 patients (135 male, 102 female; age, 70.8 ± 10.4 years) with a known primary malignancy and suspected pulmonary lesion(s) on CXR and who underwent DTS were retrospectively identified. Two radiologists (experience, 10 and 15 years) analysed in consensus CXR and DTS images and proposed a diagnosis according to a confidence score: 1 or 2 = definitely or probably benign pulmonary or extrapulmonary lesion, or pseudolesion; 3 = indeterminate; 4 or 5 = probably or definitely pulmonary lesion. DTS findings were proven by CT (n = 114 patients), CXR during follow-up (n = 105) or histology (n = 18). Final diagnoses included 77 pulmonary opacities, 26 pulmonary scars, 12 pleural lesions and 122 pulmonary pseudolesions. DTS vs CXR presented a higher (P chest radiography (CXR) in oncologic patients. • DTS improves confidence of CXR in oncologic patients. • DTS allowed avoidance of CT in about 50 % of oncologic patients.

  14. Effects of scanning resolution and digital image magnification on photostimulable phosphor imaging system

    International Nuclear Information System (INIS)

    Sakurai, Takashi; Inagaki, Masafumi; Asai, Hideomi; Koyama, Atsushi; Kashima, Isamu

    2000-01-01

    The purpose of this study is to examine the effects of changes in scanning resolution and digital magnification on the image quality and diagnostic ability of the photostimulable phosphor imaging system. Using a photostimulable phosphor imaging system, images of a human adult dried mandible phantom embedded in a 25 mm-thick epoxy resin block were made. The latent images on the photostimulable phosphor imaging plate were scanned using four different pixel sizes as follows: 25 μm x 25 μm, 50 μm x 50 μm, 100 μm x 100 μm and 200 μm x 200 μm. A primary image was produced for each pixel size. These images were also digitally magnified at powers of 2, 4 and 8 times. The gradient range, brightness and contrast of each image were adjusted to optimum levels on a cathode ray tube display, and hard copies were produced with a writing pixel size of 60 μm x 60 μm. The granularity, sharpness and anatomical diagnostic ability of the images were assessed subjectively by eight dentists. Increasing the scanning resolution tended to generally improve image quality and diagnostic ability. Visual image quality was maintained up to a pixel size of 50 μm, and diagnostic ability was maintained up to a pixel size of 100 μm. Digital image magnification degraded image quality, and more than 2-times magnification degraded diagnostic ability. Under the present experimental conditions, increasing the scanning resolution did not always lead to an improvement in image quality or diagnostic ability, and digital image magnification degraded image quality and diagnostic ability. (author)

  15. Evaluation of the Significance of Incidental Breast Lesions Detected by Chest CT

    International Nuclear Information System (INIS)

    Kim, Jae Hyun; Chang, Yun Woo; Hwang, Jung Hwa; Kim, Hyung Hwan; Lee, Eun Hye; Yang, Seung Boo

    2013-01-01

    To evaluate the significance of incidentally detected breast lesions on a chest CT scan. Thirty-six incidental breast lesions in 26 patients were detected on a chest CT scan and were correlated with breast sonography, retrospectively. Among them, twenty-four breast lesions in 20 patients that were correlated with chest CT and sonography were available to pathology or follow up sonography. The CT findings were compared with sonographic findings according to the pathologic results. Incidentally detected breast lesions on a chest CT scan were correlated with sonography in 86% (31/36). Among 24 lesions that were available to pathology or follow up sonography, seven (29.2%) lesions were malignant and 17 (70.8%) lesions were benign. CT revealed a significant difference between benign and malignant lesions in terms of shape and margin (p = 0.007; p = 0.008, respectively). The CT findings were well correlated with sonographic findings in shape and margin (p = 0.001, respectively). Incidentally detected breast lesions on chest CT can be correlated with sonography. An irregular shape or a non-circumscribed margin of breast lesions on a CT scan can be considered as a suggestive sign of malignancy.

  16. Pneumothorax size measurements on digital chest radiographs: Intra- and inter- rater reliability.

    Science.gov (United States)

    Thelle, Andreas; Gjerdevik, Miriam; Grydeland, Thomas; Skorge, Trude D; Wentzel-Larsen, Tore; Bakke, Per S

    2015-10-01

    Detailed and reliable methods may be important for discussions on the importance of pneumothorax size in clinical decision-making. Rhea's method is widely used to estimate pneumothorax size in percent based on chest X-rays (CXRs) from three measure points. Choi's addendum is used for anterioposterior projections. The aim of this study was to examine the intrarater and interrater reliability of the Rhea and Choi method using digital CXR in the ward based PACS monitors. Three physicians examined a retrospective series of 80 digital CXRs showing pneumothorax, using Rhea and Choi's method, then repeated in a random order two weeks later. We used the analysis of variance technique by Eliasziw et al. to assess the intrarater and interrater reliability in altogether 480 estimations of pneumothorax size. Estimated pneumothorax sizes ranged between 5% and 100%. The intrarater reliability coefficient was 0.98 (95% one-sided lower-limit confidence interval C 0.96), and the interrater reliability coefficient was 0.95 (95% one-sided lower-limit confidence interval 0.93). This study has shown that the Rhea and Choi method for calculating pneumothorax size has high intrarater and interrater reliability. These results are valid across gender, side of pneumothorax and whether the patient is diagnosed with primary or secondary pneumothorax. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  17. Thin chest wall is an independent risk factor for the development of pneumothorax after chest tube removal.

    Science.gov (United States)

    Anand, Rahul J; Whelan, James F; Ferrada, Paula; Duane, Therese M; Malhotra, Ajai K; Aboutanos, Michel B; Ivatury, Rao R

    2012-04-01

    The factors contributing to the development of pneumothorax after removal of chest tube thoracostomy are not fully understood. We hypothesized that development of post pull pneumothorax (PPP) after chest tube removal would be significantly lower in those patients with thicker chest walls, due to the "protective" layer of adipose tissue. All patients on our trauma service who underwent chest tube thoracostomy from July 2010 to February 2011 were retrospectively reviewed. Patient age, mechanism of trauma, and chest Abbreviated Injury Scale score were analyzed. Thoracic CTs were reviewed to ascertain chest wall thickness (CW). Thickness was measured at the level of the nipple at the midaxillary line, as perpendicular distance between skin and pleural cavity. Chest X-ray reports from immediately prior and after chest tube removal were reviewed for interval development of PPP. Data are presented as average ± standard deviation. Ninety-one chest tubes were inserted into 81 patients. Patients who died before chest tube removal (n = 11), or those without thoracic CT scans (n = 13) were excluded. PPP occurred in 29.9 per cent of chest tube removals (20/67). When PPP was encountered, repeat chest tube was necessary in 20 per cent of cases (4/20). After univariate analysis, younger age, penetrating mechanism, and thin chest wall were found to be significant risk factors for development of PPP. Chest Abbreviated Injury Scale score was similar in both groups. Logistic regression showed only chest wall thickness to be an independent risk factor for development of PPP.

  18. Atelectasis on pediatric chest CT: comparison of sedation techniques

    International Nuclear Information System (INIS)

    Sargent, M.A.; McEachern, A.M.; Jamieson, D.H.

    1999-01-01

    Background. A change in practice at our institution resulted in increased use of anesthesia for CT scan of the chest in children who required sedation. Objective. To determine whether there is a difference in the frequency or severity of pulmonary atelectasis on CT scan in children sedated by anesthesiologists compared with children sedated by radiologists using intravenous pentobarbital. Materials and methods. Retrospective blinded review of 60 CT scans of the chest performed in 41 children. Forty-one studies in children sedated by radiologists (median age 29 months) were compared with 19 studies in children sedated by anesthesiologists (median age 25 months). Results. Atelectasis sufficient to obscure pulmonary metastases was shown in 5 of 41 (12 %) radiology sedations and 13 of 19 (68 %) anesthesiology sedations (P < 0.01). Higher grades of atelectasis were recorded in children under anesthesia (P < 0.01). Conclusion. Atelectasis is more frequent and more severe in children undergoing general anesthesia compared with intravenous pentobarbital sedation. Consideration should be given to the use of forced inspiration in children anesthetized for CT scan of the chest. (orig.)

  19. A comparison of digital tomosynthesis and chest radiography in evaluating airway lesions using computed tomography as a reference

    International Nuclear Information System (INIS)

    Choo, Ji Yung; Lee, Ki Yeol; Choi, Jung Won; Yu, Ami; Kim, Je-Hyeong; Lee, Seung Heon; Kang, Eun-Young; Oh, Yu Whan

    2016-01-01

    To compare the diagnostic performance of digital tomosynthesis (DTS) and chest radiography for detecting airway abnormalities, using computed tomography (CT) as a reference. We evaluated 161 data sets from 149 patients (91 with and 70 without airway abnormalities) who had undergone radiography, DTS, and CT to detect airway problems. Radiographs and DTS were evaluated to localize and score the severity of the airway abnormalities, and to score the image quality using CT as a reference. Receiver operating characteristics (ROC), McNemar's test, weighted kappa, and the paired t-test were used for statistical analysis. The sensitivity of DTS was higher (reader 1, 93.51 %; reader 2, 94.29 %) than chest radiography (68.83 %; 71.43 %) in detecting airway lesions. The diagnostic accuracy of DTS (90.91 %; 94.70 %) was also significantly better than that of radiography (78.03 %; 82.58 %, all p < 0.05). DTS image quality was significantly better than chest radiography (1.83, 2.74; p < 0.05) in the results of both readers. The inter-observer agreement with respect to DTS findings was moderate and superior when compared to radiography findings. DTS is a more accurate and sensitive modality than radiography for detecting airway lesions that are easily obscured by soft tissue structures in the mediastinum. (orig.)

  20. Computerized detection of lung nodules in digital chest radiographs

    International Nuclear Information System (INIS)

    Giger, M.L.; Doi, K.; MacMahon, H.

    1987-01-01

    Detection of cancerous lung nodules in chest radiographs is one of the more important tasks performed by a radiologist. In addition, the ''miss rate'' associated with the radiographic detection of lung nodules is approximately 30%. A computerized scheme that alerts the radiologist to possible locations of lung nodules should allow this number of false-negative diagnoses to be reduced. The authors are developing a computer-aided nodule detection scheme based on a difference image approach. They attempt to eliminate the camouflaging background structure of the normal lung anatomy by creating, from a single-projection chest image, two images: one in which the signal-to-noise ratio (SNR) of the nodule is maximized and another in which that SNR is suppressed while the processed background remains essentially the same. Thus, the difference between these two processed images should consist of the nodule superimposed on a relatively uniform background in which the detection task may be simplified. This difference image approach is fundamentally different from conventional subtraction techniques (e.g., temporal or dual-energy subtraction) in that the two images which are subtracted arise from the same single-projection chest radiograph. Once the difference image is obtained, thresholding is performed along with tests for circularity, size and growth in order to extract the nodules. It should be noted that once an original chest image is input to the computer the nodule detection process is totally automated

  1. Lung involvement quantification in chest radiographs

    International Nuclear Information System (INIS)

    Giacomini, Guilherme; Alvarez, Matheus; Oliveira, Marcela de; Miranda, Jose Ricardo A.; Pina, Diana R.; Pereira, Paulo C.M.; Ribeiro, Sergio M.

    2014-01-01

    Tuberculosis (TB) caused by Mycobacterium tuberculosis, is an infectious disease which remains a global health problem. The chest radiography is the commonly method employed to assess the TB's evolution. The methods for quantification of abnormalities of chest are usually performed on CT scans (CT). This quantification is important to assess the TB evolution and treatment and comparing different treatments. However, precise quantification is not feasible for the amount of CT scans required. The purpose of this work is to develop a methodology for quantification of lung damage caused by TB through chest radiographs. It was developed an algorithm for computational processing of exams in Matlab, which creates a lungs' 3D representation, with compromised dilated regions inside. The quantification of lung lesions was also made for the same patients through CT scans. The measurements from the two methods were compared and resulting in strong correlation. Applying statistical Bland and Altman, all samples were within the limits of agreement, with a confidence interval of 95%. The results showed an average variation of around 13% between the two quantification methods. The results suggest the effectiveness and applicability of the method developed, providing better risk-benefit to the patient and cost-benefit ratio for the institution. (author)

  2. Quality of intensive care chest imaging

    International Nuclear Information System (INIS)

    Adam, G.; Wein, B.; Keulers, P.; Stargardt, A.; Guenther, R.W.

    1989-01-01

    The authors have evaluated the image quality of a stimulable phosphorous plate system in intensive care chest radiography. Four radiologists examined 308 chest radiographs (200 conventional, 108 digital) according to the following criteria: visibility of catheters, tubes (artificial objects), bronchi, central and peripheral vessels, diaphragm, trachea, and retrocardial lung parenchyma. Detectability of these structures was classified as good, poor, or impossible to see. In addition, optical density was measured in the region of liver, heart, and lung. Results were evaluated by Student and υ test

  3. Study on the high-frequency laser measurement of slot surface difference

    Science.gov (United States)

    Bing, Jia; Lv, Qiongying; Cao, Guohua

    2017-10-01

    In view of the measurement of the slot surface difference in the large-scale mechanical assembly process, Based on high frequency laser scanning technology and laser detection imaging principle, This paragraph designs a double galvanometer pulse laser scanning system. Laser probe scanning system architecture consists of three parts: laser ranging part, mechanical scanning part, data acquisition and processing part. The part of laser range uses high-frequency laser range finder to measure the distance information of the target shape and get a lot of point cloud data. Mechanical scanning part includes high-speed rotary table, high-speed transit and related structure design, in order to realize the whole system should be carried out in accordance with the design of scanning path on the target three-dimensional laser scanning. Data processing part mainly by FPGA hardware with LAbVIEW software to design a core, to process the point cloud data collected by the laser range finder at the high-speed and fitting calculation of point cloud data, to establish a three-dimensional model of the target, so laser scanning imaging is realized.

  4. [The best noise index combined with ASIR weighting selection in low-dose chest scanning].

    Science.gov (United States)

    Xiao, Huijuan; Hou, Ping; Liu, Jie; Gao, Jianbo; Tan, Hongna; Liang, Pan; Pu, Shi

    2015-10-06

    To discuss the best noise index combined with ASIR weighting selection in low-dose chest scanning based on BMI. 200 patients collected from May to December 2014 underwent non-contrast chest CT examinations, they were randomly assigned into standard dose group (Group A, NI15 combined with 30% ASIR) and low-dose groups (Group B, NI25 combined with 40% ASIR, Group C, NI30 combined with 50% ASIR, Group D, NI35 combined with 60% ASIR), 50 cases in each group; the patients were assigned into three groups based on BMI (kg/m2): BMI25. Signal-to-nosie ratio (SNR), contrast-to noise ratio (CNR), CT dose index volume (CTDIvol), dose-length product (DLP), effective dose (ED) and subjective scoring between the standard and low-dose groups were compared and analyzed statistically. Differences of SNR, CNR, CTDIvol, DLP and ED among groups were determined with ANOVA analysis and the consistency of diagnosis with Kappa test. SNR, CTDIvol, DLP and ED reduced with the increase of nosie index, the differences among the groups were statistically significant (P25 kg/m2 group. NI35 combined with 60% ASIR in BMIASIR in 18.5 kg/m2≤BMI≤25 kg/m2 group; NI25 combined with 40% ASIR in 18.5 kg/m2≤BMI≤25 kg/m2 group were the best parameters combination which both can significantly reduce the radiation dose and ensure the image quality.

  5. MHD-flow in slotted channels with conducting walls

    International Nuclear Information System (INIS)

    Evtushenko, I.A.; Kirillov, I.R.; Reed, C.B.

    1994-07-01

    A review of experimental results is presented for magnetohydrodynamic (MHD) flow in rectangular channels with conducting walls and high aspect ratios (longer side parallel to the applied magnetic field), which are called slotted channels. The slotted channel concept was conceived at Efremov Institute as a method for reducing MHD pressure drop in liquid metal cooled blanket design. The experiments conducted by the authors were aimed at studying both fully developed MHD-flow, and the effect of a magnetic field on the hydrodynamics of 3-D flows in slotted channels. Tests were carried out on five models of the slotted geometry. A good agreement between test and theoretical results for the pressure drop in slotted channels was demonstrated. Application of a open-quotes one-electrode movable probeclose quotes for velocity measurement permitted measurement of the M-shape velocity profiles in the slotted channels. Suppression of 3-D inertial effects in slotted channels of complex geometry was demonstrated based on potential distribution data

  6. Acceptable levels of digital image compression in chest radiology

    International Nuclear Information System (INIS)

    Smith, I.

    2000-01-01

    The introduction of picture archival and communications systems (PACS) and teleradiology has prompted an examination of techniques that optimize the storage capacity and speed of digital storage and distribution networks. The general acceptance of the move to replace conventional screen-film capture with computed radiography (CR) is an indication that clinicians within the radiology community are willing to accept images that have been 'compressed'. The question to be answered, therefore, is what level of compression is acceptable. The purpose of the present study is to provide an assessment of the ability of a group of imaging professionals to determine whether an image has been compressed. To undertake this study a single mobile chest image, selected for the presence of some subtle pathology in the form of a number of septal lines in both costphrenic angles, was compressed to levels of 10:1, 20:1 and 30:1. These images were randomly ordered and shown to the observers for interpretation. Analysis of the responses indicates that in general it was not possible to distinguish the original image from its compressed counterparts. Furthermore, a preference appeared to be shown for images that have undergone low levels of compression. This preference can most likely be attributed to the 'de-noising' effect of the compression algorithm at low levels. Copyright (1999) Blackwell Science Pty. Ltd

  7. Pumping slots: impedances and power losses

    Energy Technology Data Exchange (ETDEWEB)

    Kurennoy, S [Maryland Univ., College Park, MD (United States). Dept. of Physics

    1996-08-01

    Contributions of pumping slots to the beam coupling impedances and power losses in a B-factory ring are considered. While their leading contribution is to the inductive impedance, for high-intensity machines with short bunches like e{sup +}e{sup -} B-factories the real part of the impedance and related loss factors are also important. Using an analytical approach we calculate the coupling impedances and loss factors due to slots in a ring with an arbitrary cross section of the vacuum chamber. Effects of the slot tilt on the beam impedance are also considered, and restrictions on the tilt angle are derived from limitations on the impedance increase. The power leakage through the slots is discussed briefly. The results are applied to the KEK B-factory. (author)

  8. Digital radiographic scanning installation with multiwire proportional chamber

    International Nuclear Information System (INIS)

    Babichev, E.A.; Baru, S.E.; Khabakhpashev, A.G.; Kolachev, G.M.; Savinov, G.A.; Shekhtman, L.I.; Sidorov, V.A.; Volobuev, A.I.

    1991-01-01

    A digital radiographic installation for medical diagnostics is described. Distribution of X-ray intensity in horizontal direction is measured by a one-coordinate MWPC with a 1.2 mm channel. A two-dimensional image is produced by mechanical scanning with 1 mm channel width in vertical direction. The number of pixels in the image is 320x256. The MWPC works in count mode with quantum efficiency 25% (50 keV) and count rate capability 600 kHz/channel. Results of testing a prototype with 64 channels 0.6 mm width are also presented. (orig.)

  9. Digital radiography of scoliosis with a scanning method: radiation dose optimization

    Energy Technology Data Exchange (ETDEWEB)

    Geijer, Haakan; Andersson, Torbjoern [Department of Radiology, Oerebro University Hospital, 701 85 Oerebro (Sweden); Verdonck, Bert [Philips Medical Systems, P.O. Box 10,000, 5680 Best (Netherlands); Beckman, Karl-Wilhelm; Persliden, Jan [Department of Medical Physics, Oerebro University Hospital, 701 85 Oerebro (Sweden)

    2003-03-01

    The aim of this study was optimization of the radiation dose-image quality relationship for a digital scanning method of scoliosis radiography. The examination is performed as a digital multi-image translation scan that is reconstructed to a single image in a workstation. Entrance dose was recorded with thermoluminescent dosimeters placed dorsally on an Alderson phantom. At the same time, kerma area product (KAP) values were recorded. A Monte Carlo calculation of effective dose was also made. Image quality was evaluated with a contrast-detail phantom and Visual Grading. The radiation dose was reduced by lowering the image intensifier entrance dose request, adjusting pulse frequency and scan speed, and by raising tube voltage. The calculated effective dose was reduced from 0.15 to 0.05 mSv with reduction of KAP from 1.07 to 0.25 Gy cm{sup 2} and entrance dose from 0.90 to 0.21 mGy. The image quality was reduced with the Image Quality Figure going from 52 to 62 and a corresponding reduction in image quality as assessed with Visual Grading. The optimization resulted in a dose reduction to 31% of the original effective dose with an acceptable reduction in image quality considering the intended use of the images for angle measurements. (orig.)

  10. Breast dose reduction for chest CT by modifying the scanning parameters based on the pre-scan size-specific dose estimate (SSDE)

    Energy Technology Data Exchange (ETDEWEB)

    Kidoh, Masafumi; Utsunomiya, Daisuke; Oda, Seitaro; Nakaura, Takeshi; Yuki, Hideaki; Hirata, Kenichiro; Namimoto, Tomohiro; Sakabe, Daisuke; Hatemura, Masahiro; Yamashita, Yasuyuki [Kumamoto University, Department of Diagnostic Radiology, Faculty of Life Sciences, Honjo, Kumamoto (Japan); Funama, Yoshinori [Kumamoto University, Department of Medical Physics, Faculty of Life Sciences, Honjo, Kumamoto (Japan)

    2017-06-15

    To investigate the usefulness of modifying scanning parameters based on the size-specific dose estimate (SSDE) for a breast-dose reduction for chest CT. We scanned 26 women with a fixed volume CT dose index (CTDI{sub vol}) (15 mGy) and another 26 with a fixed SSDE (15 mGy) protocol (protocol 1 and 2, respectively). In protocol 2, tube current was calculated based on the patient habitus obtained on scout images. We compared the mean breast dose and the inter-patient breast dose variability and performed linear regression analysis of the breast dose and the body mass index (BMI) of the two protocols. The mean breast dose was about 35 % lower under protocol 2 than protocol 1 (10.9 mGy vs. 16.8 mGy, p < 0.01). The inter-patient breast dose variability was significantly lower under protocol 2 than 1 (1.2 mGy vs. 2.5 mGy, p < 0.01). We observed a moderate negative correlation between the breast dose and the BMI under protocol 1 (r = 0.43, p < 0.01); there was no significant correlation (r = 0.06, p = 0.35) under protocol 2. The SSDE-based protocol achieved a reduction in breast dose and in inter-patient breast dose variability. (orig.)

  11. ODK Scan: Digitizing Data Collection and Impacting Data Management Processes in Pakistan’s Tuberculosis Control Program

    Directory of Open Access Journals (Sweden)

    Syed Mustafa Ali

    2016-10-01

    Full Text Available The present grievous tuberculosis situation can be improved by efficient case management and timely follow-up evaluations. With the advent of digital technology, this can be achieved through quick summarization of the patient-centric data. The aim of our study was to assess the effectiveness of the ODK Scan paper-to-digital system during a testing period of three months. A sequential, explanatory mixed-method research approach was employed to elucidate technology use. Training, smartphones, the application and 3G-enabled SIMs were provided to the four field workers. At the beginning, baseline measures of the data management aspects were recorded and compared with endline measures to determine the impact of ODK Scan. Additionally, at the end of the study, users’ feedback was collected regarding app usability, user interface design and workflow changes. A total of 122 patients’ records were retrieved from the server and analysed in terms of quality. It was found that ODK Scan recognized 99.2% of multiple choice fill-in bubble responses and 79.4% of numerical digit responses correctly. However, the overall quality of the digital data was decreased in comparison to manually entered data. Using ODK Scan, a significant time reduction is observed in data aggregation and data transfer activities, but data verification and form-filling activities took more time. Interviews revealed that field workers saw value in using ODK Scan, but they were more concerned about the time-consuming aspects of the use of ODK Scan. Therefore, it is concluded that minimal disturbance in the existing workflow, continuous feedback and value additions are the important considerations for the implementing organization to ensure technology adoption and workflow improvements.

  12. A scanning electron microscopic investigation of ceramic orthodontic brackets

    International Nuclear Information System (INIS)

    McDonald, F.; Toms, A.P.

    1990-01-01

    Ceramic brackets were introduced to overcome the esthetic disadvantages of stainless steel brackets. The clinical impression of these brackets is very favorable. However, the sliding mechanics used in the Straightwire (A Company, San Diego, CA, USA) system appear to produce slower tooth movements with ceramic compared to stainless steel brackets. To determine whether this was due to any obvious mechanical problem in the bracket slot, Transcend (Unitek Corporation/3M, Monrovia, CA, USA) ceramic brackets were examined by a scanning electron microscope and compared to stainless steel brackets.Consistently, large surface defects were found in the ceramic bracket slots that were not present in the metal bracket slots. These irregularities could obviously hinder the sliding mechanics of the bracket slot-archwire system and create a greater demand on anchorage. Conversely, the fitting surface of the Transcend ceramic bracket showed extremely smooth surface characteristics, and it would seem advisable for the manufacturers to incorporate this surface within the bracket slot. (author)

  13. Student-Led Objective Tutorial (SLOT) in Medical Education.

    Science.gov (United States)

    Sivagnanam, Gurusamy; Saraswathi, Simansalam; Rajasekaran, Aiyalu

    2006-12-01

    Purpose - To assess an innovative tutoring program named 'Student-Led Objective Tutorial' (SLOT) among undergraduate medical students. Method - The program was conceptualized by the Pharmacology Unit of Faculty of Medicine and Health Sciences, Asian Institute of Medicine Science & Technology (AIMST), Malaysia and implemented in the middle of 2005. A cohort of 246 medical undergraduate students (spread across 5 consecutive batches) participated. Following a brief explanation on the purpose and nature of SLOT, each batch was divided into small groups and was given a reading assignment on 4 previously delivered lecture topics. Each group was asked to prepare 3-5 multiple choice questions (MCQs) of their own in PowerPoint format to be presented, in turns, to the whole class on the day of SLOT. The proceedings were facilitated by 2 lecturers. Student feedback on the efficacy and benefits were assessed through an anonymous self administered questionnaire. Results - About 76% (188) of the students favored SLOT. The acceptance rate of SLOT was higher among males. There was no significant difference between batches in their opinions on whether to pursue SLOT in future. The most prevalent positive comment was that SLOT enhanced learning skills, and the negative comment being, it consumed more time. Conclusions - SLOT is a novel tutorial method which can offset faculty shortage with advantages like enhanced interest among teachers and learners, uniform reach of content, opportunities for group learning, and involvement of visual aids as teaching-learning (T-L) method. SLOT unraveled the students' potential of peer tutoring both inside as well as outside the classroom. Consumer tutors (students) can be tapped as a resource for SLOT for all subjects and courses in healthcare teaching.

  14. Synchronous digitization for high dynamic range lock-in amplification in beam-scanning microscopy

    Energy Technology Data Exchange (ETDEWEB)

    Muir, Ryan D.; Sullivan, Shane Z.; Oglesbee, Robert A.; Simpson, Garth J., E-mail: gsimpson@purdue.edu [Department of Chemistry, Purdue University, 560 Oval Drive, West Lafayette, Indiana 47907 (United States)

    2014-03-15

    Digital lock-in amplification (LIA) with synchronous digitization (SD) is shown to provide significant signal to noise (S/N) and linear dynamic range advantages in beam-scanning microscopy measurements using pulsed laser sources. Direct comparisons between SD-LIA and conventional LIA in homodyne second harmonic generation measurements resulted in S/N enhancements consistent with theoretical models. SD-LIA provided notably larger S/N enhancements in the limit of low light intensities, through the smooth transition between photon counting and signal averaging developed in previous work. Rapid beam scanning instrumentation with up to video rate acquisition speeds minimized photo-induced sample damage. The corresponding increased allowance for higher laser power without sample damage is advantageous for increasing the observed signal content.

  15. 1024 matrix image reconstruction: usefulness in high resolution chest CT

    International Nuclear Information System (INIS)

    Jeong, Sun Young; Chung, Myung Jin; Chong, Se Min; Sung, Yon Mi; Lee, Kyung Soo

    2006-01-01

    We tried to evaluate whether high resolution chest CT with a 1,024 matrix has a significant advantage in image quality compared to a 512 matrix. Each set of 512 and 1024 matrix high resolution chest CT scans with both 0.625 mm and 1.25 mm slice thickness were obtained from 26 patients. Seventy locations that contained twenty-four low density lesions without sharp boundary such as emphysema, and forty-six sharp linear densities such as linear fibrosis were selected; these were randomly displayed on a five mega pixel LCD monitor. All the images were masked for information concerning the matrix size and slice thickness. Two chest radiologists scored the image quality of each ar rowed lesion as follows: (1) undistinguishable, (2) poorly distinguishable, (3) fairly distinguishable, (4) well visible and (5) excellently visible. The scores were compared from the aspects of matrix size, slice thickness and the different observers by using ANOVA tests. The average and standard deviation of image quality were 3.09 (± .92) for the 0.625 mm x 512 matrix, 3.16 (± .84) for the 0.625 mm x 1024 matrix, 2.49 (± 1.02) for the 1.25 mm x 512 matrix, and 2.35 (± 1.02) for the 1.25 mm x 1024 matrix, respectively. The image quality on both matrices of the high resolution chest CT scans with a 0.625 mm slice thickness was significantly better than that on the 1.25 mm slice thickness (ρ < 0.001). However, the image quality on the 1024 matrix high resolution chest CT scans was not significantly different from that on the 512 matrix high resolution chest CT scans (ρ = 0.678). The interobserver variation between the two observers was not significant (ρ = 0.691). We think that 1024 matrix image reconstruction for high resolution chest CT may not be clinical useful

  16. Chest imaging in aids - radiological findings with pathologic correlation: review article

    International Nuclear Information System (INIS)

    Qazi, A.S.

    1999-01-01

    Majority of life threatening illnesses in AIDS begin as pulmonary infections and a radiologist must always seriously consider the possibility of HIV infection and its manifestation when confronting an abnormal chest study in a young adult. Chest radiography may be normal in up to 15% of patients with proven pulmonary involvement or the radiographic picture may be confusing due to atypical appearances of opportunistic infections in immuno-compromised host, compounded further by concomitant appearance of neoplastic complications like Kaposi AIDS relate lymphoma. Cases with normal chest radiograph but high degree of suspicion of chest disease need to be evaluated by CT scan which has been found to be superior to chest radiography in identifying patient with and without chest disease and in the differential diagnosis of pulmonary complications in patients with AIDS. Radio nuclear scans and MRI have some role only in selected few cases. Combining imaging features with clinical presentation, CD4 lymphocyte count, previous treatment and underlying risk group can narrow down differential diagnosis, expedite treatment and may be helpful in preventing complications. (author)

  17. Management of chest drainage tubes after lung surgery.

    Science.gov (United States)

    Satoh, Yukitoshi

    2016-06-01

    Since chest tubes have been routinely used to drain the pleural space, particularly after lung surgery, the management of chest tubes is considered to be essential for the thoracic surgeon. The pleural drainage system requires effective drainage, suction, and water-sealing. Another key point of chest tube management is that a water seal is considered to be superior to suction for most air leaks. Nowadays, the most common pleural drainage device attached to the chest tube is the three-bottle system. An electronic chest drainage system has been developed that is effective in standardizing the postoperative management of chest tubes. More liberal use of digital drainage devices in the postoperative management of the pleural space is warranted. The removal of chest tubes is a common procedure occurring almost daily in hospitals throughout the world. Extraction of the tube is usually done at the end of full inspiration or at the end of full expiration. The tube removal technique is not as important as how it is done and the preparation for the procedure. The management of chest tubes must be based on careful observation, the patient's characteristics, and the operative procedures that had been performed.

  18. Chest computed tomography in children undergoing extra-corporeal membrane oxygenation: a 9-year single-centre experience

    Energy Technology Data Exchange (ETDEWEB)

    Goodwin, Susie J.; Calder, Alistair D. [Great Ormond Street Hospital for Children NHS Foundation Trust, Radiology Department, London (United Kingdom); Randle, Elise; Iguchi, Akane; Brown, Katherine; Hoskote, Aparna [Great Ormond Street Hospital for Children NHS Foundation Trust, Cardiac Intensive Care and ECMO, London (United Kingdom)

    2014-06-15

    We retrospectively reviewed the imaging findings, indications, technique and clinical impact in children who had undergone chest CT while undergoing extra-corporeal membrane oxygenation (ECMO). Radiology and ECMO databases were searched to identify all 19 children who had undergone chest CT (20 scans in total) while on ECMO at our institution between May 2003 and May 2012. We reviewed all CT scans for imaging findings. Chest CT is performed in a minority of children on ECMO (4.5% in our series). Timing of chest CT following commencement of ECMO varied among patient groups but generally it was performed earlier in the neonatal group. Clinically significant imaging findings were found in the majority of chest CT scans. Many scans contained several findings, with most cases demonstrating parenchymal or pleural abnormalities. Case examples illustrate the spectrum of imaging findings, including underlying pathology such as necrotising pneumonia and severe barotrauma, and ECMO-related complications such as tension haemothoraces and cannula migration. The results of chest CT led to a change in patient management in 16 of 19 children (84%). There were no adverse events related to patient transfer. An understanding of scan technique and awareness of potential findings is important for the radiologist to provide prompt and optimal image acquisition and interpretation in appropriate patients. (orig.)

  19. T-scan III system diagnostic tool for digital occlusal analysis in orthodontics - a modern approach.

    Science.gov (United States)

    Trpevska, Vesna; Kovacevska, Gordana; Benedeti, Alberto; Jordanov, Bozidar

    2014-01-01

    This systematic literature review was performed to establish the mechanism, methodology, characteristics, clinical application and opportunities of the T-Scan III System as a diagnostic tool for digital occlusal analysis in different fields of dentistry, precisely in orthodontics. Searching of electronic databases, using MEDLINE and PubMed, hand searching of relevant key journals, and screening of reference lists of included studies with no language restriction was performed. Publications providing statistically examined data were included for systematic review. Twenty potentially relevant Randomized Controlled Trials (RCTs) were identified. Only ten met the inclusion criteria. The literature demonstrates that using digital occlusal analysis with T-Scan III System in orthodontics has significant advantage with regard to the capability of measuring occlusal parameters in static positions and during dynamic of the mandible. Within the scope of this systematic review, there is evidence to support that T-Scan system is rapid and accurate in identifying the distribution of the tooth contacts and it shows great promise as a clinical diagnostic screening device for occlusion and for improving the occlusion after various dental treatments. Additional clinical studies are required to advance the indication filed of this system. Importance of using digital occlusal T-Scan analysis in orthodontics deserves further investigation.

  20. Evaluation of surface roughness of the bracket slot floor—a 3D perspective study

    Directory of Open Access Journals (Sweden)

    Chetankumar O. Agarwal

    2016-01-01

    Full Text Available Abstract Background An important constituent of an orthodontic appliance is orthodontic brackets. It is either the bracket or the archwire that slides through the bracket slot, during sliding mechanics. Overcoming the friction between the two surfaces demands an important consideration in an appliance design. The present study investigated the surface roughness of four different commercially available stainless steel brackets. Methods All tests were carried out to analyse quantitatively the morphological surface of the bracket slot floor with the help of scanning electron microscope (SEM machine and to qualitatively analyse the average surface roughness (Sa of the bracket slot floor with the help of a three-dimensional (3D non-contact optical surface profilometer machine. Results The SEM microphotographs were evaluated with the help of visual analogue scale, the surface roughness for group A = 0—very rough surface, group C = 1—rough surface, group B = 2—smooth surface, and group D = 3—very smooth surface. Surface roughness evaluation with the 3D non-contact optical surface profilometer machine was highest for group A, followed by group C, group B and group D. Groups B and D provided smooth surface roughness; however, group D had the very smooth surface with values 0.74 and 0.75 for mesial and distal slots, respectively. Conclusions Evaluation of surface roughness of the bracket slot floor with both SEM and profilometer machine led to the conclusion that the average surface roughness was highest for group A, followed by group C, group B and group D.

  1. Slotting allowances to coordinate manufacturers’ retail sales effort

    OpenAIRE

    Foros, Øystein; Kind, Hans Jarle; Sand, Jan Yngve

    2007-01-01

    Slotting allowances are fees paid by manufacturers to get access to retailers’ shelf space. Although the main attention towards slotting allowances has been within the grocery industry, slotting allowances have also been applied within e.g. e-commerce and mobile telephony. In these industries we observe that distributors have large market power due to their control of access to customers. We analyse how shifting bargaining power from manufacturers to retailers and the use of slotting allowanc...

  2. Patient-specific radiation dose and cancer risk for pediatric chest CT.

    Science.gov (United States)

    Li, Xiang; Samei, Ehsan; Segars, W Paul; Sturgeon, Gregory M; Colsher, James G; Frush, Donald P

    2011-06-01

    To estimate patient-specific radiation dose and cancer risk for pediatric chest computed tomography (CT) and to evaluate factors affecting dose and risk, including patient size, patient age, and scanning parameters. The institutional review board approved this study and waived informed consent. This study was HIPAA compliant. The study included 30 patients (0-16 years old), for whom full-body computer models were recently created from clinical CT data. A validated Monte Carlo program was used to estimate organ dose from eight chest protocols, representing clinically relevant combinations of bow tie filter, collimation, pitch, and tube potential. Organ dose was used to calculate effective dose and risk index (an index of total cancer incidence risk). The dose and risk estimates before and after normalization by volume-weighted CT dose index (CTDI(vol)) or dose-length product (DLP) were correlated with patient size and age. The effect of each scanning parameter was studied. Organ dose normalized by tube current-time product or CTDI(vol) decreased exponentially with increasing average chest diameter. Effective dose normalized by tube current-time product or DLP decreased exponentially with increasing chest diameter. Chest diameter was a stronger predictor of dose than weight and total scan length. Risk index normalized by tube current-time product or DLP decreased exponentially with both chest diameter and age. When normalized by DLP, effective dose and risk index were independent of collimation, pitch, and tube potential (chest CT protocols. http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.11101900/-/DC1. RSNA, 2011

  3. New developments in ultrasonic imaging of the chest and other body organs

    International Nuclear Information System (INIS)

    Campbell, G.W.; Anderson, A.L.

    1978-01-01

    The ultrasonic imaging system described herein was developed to measure chest-wall thickness and the percentage of fat in the chest and around other body organs. The system uses pulse-echo techniques to transmit and detect sound waves reflected from the interfaces of body organs and adjacent tissue. A computer draws these interfaces on color scans, and a code is used to exponentially average data from several points on each scan to find the average thicknesses of the chest wall and fat layers. These average thicknesses are then used to adjust x-ray calibration factors for plutonium lung counters. The correction factor for three subjects measured for fat content ranging from 12.6 to 22.2% was 18 to 41%. The ultrasonic system also defines the shape and position of the kidneys and liver so we are able to more accurately place detectors on the body during in-vivo radiation measurements. We have also developed a technique for displaying the interfaces from a series of ultrasonic chest scans to produce a topographical map that enables us to better understand the shape and contour of the lung and chest-wall interface

  4. Chest trauma in children: current imaging guidelines and techniques.

    LENUS (Irish Health Repository)

    Moore, Michael A

    2011-09-01

    Given the heterogeneous nature of pediatric chest trauma, the optimal imaging approach is tailored to the specific patient. Chest radiography remains the most important imaging modality for initial triage. The decision to perform a chest computed tomography scan should be based on the nature of the trauma, the child\\'s clinical condition, and the initial radiographic findings, taking the age-related pretest probabilities of serious injury into account. The principles of as low as reasonably achievable and Image Gently should be followed. The epidemiology and pathophysiology, imaging techniques, characteristic findings, and evidence-based algorithms for pediatric chest trauma are discussed.

  5. Chest X-Ray (Chest Radiography)

    Science.gov (United States)

    ... Resources Professions Site Index A-Z X-ray (Radiography) - Chest Chest x-ray uses a very small dose ... Radiography? What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic ...

  6. Development of Portable Digital Radiography System with a Device for Monitoring X-ray Source-Detector Angle and Its Application in Chest Imaging

    Directory of Open Access Journals (Sweden)

    Tae-Hoon Kim

    2017-03-01

    Full Text Available This study developed a device measuring the X-ray source-detector angle (SDA and evaluated the imaging performance for diagnosing chest images. The SDA device consisted of Arduino, an accelerometer and gyro sensor, and a Bluetooth module. The SDA values were compared with the values of a digital angle meter. The performance of the portable digital radiography (PDR was evaluated using the signal-to-noise (SNR, contrast-to-noise ratio (CNR, spatial resolution, distortion and entrance surface dose (ESD. According to different angle degrees, five anatomical landmarks were assessed using a five-point scale. The mean SNR and CNR were 182.47 and 141.43. The spatial resolution and ESD were 3.17 lp/mm (157 μm and 0.266 mGy. The angle values of the SDA device were not significantly difference as compared to those of the digital angle meter. In chest imaging, the SNR and CNR values were not significantly different according to the different angle degrees. The visibility scores of the border of the heart, the fifth rib and the scapula showed significant differences according to different angles (p < 0.05, whereas the scores of the clavicle and first rib were not significant. It is noticeable that the increase in the SDA degree was consistent with the increases of the distortion and visibility score. The proposed PDR with a SDA device would be useful for application in the clinical radiography setting according to the standard radiography guidelines.

  7. The forgotten view: Chest X-ray - Lateral view

    Directory of Open Access Journals (Sweden)

    Abraham M. Ittyachen

    2017-01-01

    Full Text Available With CT (computed tomography chest gaining more importance as a diagnostic tool, chest X-ray especially the lateral view is taken less commonly nowadays. Besides CT chest is also proven to be superior to chest X-ray in patients with major blunt trauma. We are presenting a 68-year old male who was partially treated from outside for a left sided pneumonia. He came to our hospital because of persisting chest pain. Chest X-ray, frontal view (postero-anterior was almost normal except for a mild opacity in the left lower zone. CT scan of the chest revealed a fluid collection posteriorly enclosed within enhancing pleura. Chest X-ray, left lateral view showed a corresponding posterior pleural based opacity. We are presenting this case to highlight the importance of the lateral view of the chest X-ray. In selected cases there is still a role for the lateral view. With the three dimensional visualization provided by the CT, the lateral view of the chest may be easier to understand. Consequent to the initial diagnosis by CT further follow up can be done with the chest X-ray. In a limited way this mitigates unnecessary expenditure and more importantly prevents the patient from exposure to harmful radiation in the form of repeated CT.

  8. Pulmonary disease in cystic fibrosis: assessment with chest CT at chest radiography dose levels.

    Science.gov (United States)

    Ernst, Caroline W; Basten, Ines A; Ilsen, Bart; Buls, Nico; Van Gompel, Gert; De Wachter, Elke; Nieboer, Koenraad H; Verhelle, Filip; Malfroot, Anne; Coomans, Danny; De Maeseneer, Michel; de Mey, Johan

    2014-11-01

    To investigate a computed tomographic (CT) protocol with iterative reconstruction at conventional radiography dose levels for the assessment of structural lung abnormalities in patients with cystic fibrosis ( CF cystic fibrosis ). In this institutional review board-approved study, 38 patients with CF cystic fibrosis (age range, 6-58 years; 21 patients 18 years) underwent investigative CT (at minimal exposure settings combined with iterative reconstruction) as a replacement of yearly follow-up posteroanterior chest radiography. Verbal informed consent was obtained from all patients or their parents. CT images were randomized and rated independently by two radiologists with use of the Bhalla scoring system. In addition, mosaic perfusion was evaluated. As reference, the previous available conventional chest CT scan was used. Differences in Bhalla scores were assessed with the χ(2) test and intraclass correlation coefficients ( ICC intraclass correlation coefficient s). Radiation doses for CT and radiography were assessed for adults (>18 years) and children (chest CT protocol can replace the two yearly follow-up chest radiographic examinations without major dose penalty and with similar diagnostic quality compared with conventional CT.

  9. Spiral Slotted Microstrip Antenna Design for 700 MHz Band Application

    Directory of Open Access Journals (Sweden)

    Ricardo Meneses González

    2016-01-01

    Full Text Available This work describes the design and implementation of spiral slotted microstrip antenna. Recently, just like other countries, in Mexico, terrestrial digital television has been implemented (analogic shutdown; as a consequence, the 700 MHz UFH Band (698–806 MHz has been opened to new telecommunications services, particularly wireless mobile communication. This technological advance represents a radio mobile antenna design challenge because it is necessary to design an antenna whose dimensions must be small enough, which satisfies gain, resonance frequency, and bandwidth requirements and is of low cost.

  10. Digitally controlled analog proportional-integral-derivative (PID) controller for high-speed scanning probe microscopy

    Science.gov (United States)

    Dukic, Maja; Todorov, Vencislav; Andany, Santiago; Nievergelt, Adrian P.; Yang, Chen; Hosseini, Nahid; Fantner, Georg E.

    2017-12-01

    Nearly all scanning probe microscopes (SPMs) contain a feedback controller, which is used to move the scanner in the direction of the z-axis in order to maintain a constant setpoint based on the tip-sample interaction. The most frequently used feedback controller in SPMs is the proportional-integral (PI) controller. The bandwidth of the PI controller presents one of the speed limiting factors in high-speed SPMs, where higher bandwidths enable faster scanning speeds and higher imaging resolution. Most SPM systems use digital signal processor-based PI feedback controllers, which require analog-to-digital and digital-to-analog converters. These converters introduce additional feedback delays which limit the achievable imaging speed and resolution. In this paper, we present a digitally controlled analog proportional-integral-derivative (PID) controller. The controller implementation allows tunability of the PID gains over a large amplification and frequency range, while also providing precise control of the system and reproducibility of the gain parameters. By using the analog PID controller, we were able to perform successful atomic force microscopy imaging of a standard silicon calibration grating at line rates up to several kHz.

  11. Stereoscopic display in a slot machine

    Science.gov (United States)

    Laakso, M.

    2012-03-01

    This paper reports the results of a user trial with a slot machine equipped with a stereoscopic display. The main research question was to find out what kind of added value does stereoscopic 3D (S-3D) bring to slot games? After a thorough literature survey, a novel gaming platform was designed and implemented. Existing multi-game slot machine "Nova" was converted to "3DNova" by replacing the monitor with an S-3D display and converting six original games to S-3D format. To evaluate the system, several 3DNova machines were put available for players for four months. Both qualitative and quantitative analysis was carried out from statistical values, questionnaires and observations. According to the results, people find the S-3D concept interesting but the technology is not optimal yet. Young adults and adults were fascinated by the system, older people were more cautious. Especially the need to wear stereoscopic glasses provide a challenge; ultimate system would probably use autostereoscopic technology. Also the games should be designed to utilize its full power. The main contributions of this paper are lessons learned from creating an S-3D slot machine platform and novel information about human factors related to stereoscopic slot machine gaming.

  12. Dual Mode Slotted Monopole Antenna

    Science.gov (United States)

    2017-01-05

    May 2017 The below identified patent application is available for licensing. Requests for information should be addressed to: TECHNOLOGY...REFERENCE TO OTHER PATENT APPLICATIONS [0002] None. BACKGROUND OF THE INVENTION (1) Field of the Invention [0003] The present invention is directed...to connect across slot 14. [0005] In United States Patent No. 6,127,983 to Rivera and Josypenko, it has been shown that the slotted cylinder

  13. Comparison of conventional study model measurements and 3D digital study model measurements from laser scanned dental impressions

    Science.gov (United States)

    Nugrahani, F.; Jazaldi, F.; Noerhadi, N. A. I.

    2017-08-01

    The field of orthodontics is always evolving,and this includes the use of innovative technology. One type of orthodontic technology is the development of three-dimensional (3D) digital study models that replace conventional study models made by stone. This study aims to compare the mesio-distal teeth width, intercanine width, and intermolar width measurements between a 3D digital study model and a conventional study model. Twelve sets of upper arch dental impressions were taken from subjects with non-crowding teeth. The impressions were taken twice, once with alginate and once with polivinylsiloxane. The alginate impressions used in the conventional study model and the polivinylsiloxane impressions were scanned to obtain the 3D digital study model. Scanning was performed using a laser triangulation scanner device assembled by the School of Electrical Engineering and Informatics at the Institut Teknologi Bandung and David Laser Scan software. For the conventional model, themesio-distal width, intercanine width, and intermolar width were measured using digital calipers; in the 3D digital study model they were measured using software. There were no significant differences between the mesio-distal width, intercanine width, and intermolar width measurments between the conventional and 3D digital study models (p>0.05). Thus, measurements using 3D digital study models are as accurate as those obtained from conventional study models

  14. Method for dose-reduced 3D catheter tracking on a scanning-beam digital x-ray system using dynamic electronic collimation

    Science.gov (United States)

    Dunkerley, David A. P.; Funk, Tobias; Speidel, Michael A.

    2016-03-01

    Scanning-beam digital x-ray (SBDX) is an inverse geometry x-ray fluoroscopy system capable of tomosynthesis-based 3D catheter tracking. This work proposes a method of dose-reduced 3D tracking using dynamic electronic collimation (DEC) of the SBDX scanning x-ray tube. Positions in the 2D focal spot array are selectively activated to create a regionof- interest (ROI) x-ray field around the tracked catheter. The ROI position is updated for each frame based on a motion vector calculated from the two most recent 3D tracking results. The technique was evaluated with SBDX data acquired as a catheter tip inside a chest phantom was pulled along a 3D trajectory. DEC scans were retrospectively generated from the detector images stored for each focal spot position. DEC imaging of a catheter tip in a volume measuring 11.4 cm across at isocenter required 340 active focal spots per frame, versus 4473 spots in full-FOV mode. The dose-area-product (DAP) and peak skin dose (PSD) for DEC versus full field-of-view (FOV) scanning were calculated using an SBDX Monte Carlo simulation code. DAP was reduced to 7.4% to 8.4% of the full-FOV value, consistent with the relative number of active focal spots (7.6%). For image sequences with a moving catheter, PSD was 33.6% to 34.8% of the full-FOV value. The root-mean-squared-deviation between DEC-based 3D tracking coordinates and full-FOV 3D tracking coordinates was less than 0.1 mm. The 3D distance between the tracked tip and the sheath centerline averaged 0.75 mm. Dynamic electronic collimation can reduce dose with minimal change in tracking performance.

  15. S-parameters for weakly excited slots

    DEFF Research Database (Denmark)

    Albertsen, Niels Christian

    1999-01-01

    A simple approach to account for parasitic effects in weakly excited slots cut in the broad wall of a rectangular waveguide is proposed......A simple approach to account for parasitic effects in weakly excited slots cut in the broad wall of a rectangular waveguide is proposed...

  16. Slot Machine Response Frequency Predicts Pathological Gambling

    DEFF Research Database (Denmark)

    Linnet, Jakob; Rømer Thomsen, Kristine; Møller, Arne

    2013-01-01

    Slot machines are among the most addictive forms of gambling, and pathological gambling slot machine players represent the largest group of treatment seekers, accounting for 35% to 93% of the population. Pathological gambling sufferers have significantly higher response frequency (games / time......) on slot machines compared with non-problem gamblers, which may suggest increased reinforcement of the gambling behavior in pathological gambling. However, to date it is unknown whether or not the increased response frequency in pathological gambling is associated with symptom severity of the disorder....... This study tested the hypothesis that response frequency is associated with symptom severity in pathological gambling. We tested response frequency among twenty-two pathological gambling sufferers and twenty-one non-problem gamblers on a commercially available slot machine, and screened for pathological...

  17. Role of computed tomography in blunt chest trauma

    International Nuclear Information System (INIS)

    Cho, Jae Hyun; Kim, Sang Jin; Lee, Chan Wha; Kim, Hae Kyoon

    1994-01-01

    In patient with blunt trauma of chest, supine AP x-ray cannot differentiate the lung contusion, laceration, atelectasis, and hemothorax definitely. Therefore, computed tomographic evaluation is needed for accurate evaluation of the injuries. In our knowledge, there are few reports about CT findings of blunt chest trauma, in our country, therefore we tried to fluid the characteristic CT findings in patients with blunt trauma. We analyzed the plain x-ray and CT image of 4 patients with blunt chest trauma. Location and morphology of lung parenchymal contusion and laceration, hemopneumothorax, chest wall injuries and location of chest tube. Lung parenchymal contusion was noted in 53 segments., of 16 patients infiltration(n=27 segment), and multiple nodular pattern was noted in 15 segment, pattern of consolidation along the lung periphery was seen in 11 segment. Laceration was noted in 18 lesion and most commonly located in paravertebral area(b=8). CT scan of chest in patient with blunt chest trauma, provides accurate information of the pattern of injuries, and localization, therefore, should be performed as possible

  18. Dose profile in computed tomography chest scan; Perfil de dose em varredura de torax em tomografia computadorizada

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Bruno B., E-mail: bbo@cdtn.b [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil). Pos-graduacao em Ciencias e Tecnologia das Radiacoes, Minerais e Materiais; Mourao, Arnaldo P. [Centro Federal de Educacao Tecnologica de Minas Gerais (CEFET/MG), Belo Horizonte, MG (Brazil); Alonso, Thessa C.; Silva, Teogenes A. da [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2010-07-01

    For the optimization of the patient dose in computed tomography (CT), the Brazilian legislation only established the diagnostic reference levels (DRL's) in terms of Multiple Scan Average Dose (MSAD) in a typical adult as a parameter of quality control of CT scanners. Conformity to the DRL's can be verified by measuring the dose distribution in CT scans and MSAD determination. An analysis of the quality of CT scans of the metropolitan region of Belo Horizonte is necessary by conducting pertinent tests to the study that are presented in the ANVISA (National Agency of Sanitary Vigilance) Guide. The purpose of this study is to investigate, in a chest scan, the variation of dose in CT. To measure the dose profile are used lithium fluoride thermoluminescent dosimeters (TLD-100 Rod) distributed in cylinders positioned in peripheral and central regions of a phantom of polymethylmethacrylate (PMMA). The data obtained allow us to observe the variation of the dose profile inside the phantom. The peripheral region shows higher dose values than the central region. The longitudinal variation can be observed and the maximum dose was recorded at the edges of the phantom (41,58{+-}5,10) mGy at the midpoint of the longitudinal axis. The results will contribute to disseminate the proper procedure and optimize the dosimetry and the tests of quality control in CT, as well as make a critical analysis of the DRL's. (author)

  19. Digitally generated excitation and near-baseband quadrature detection of rapid scan EPR signals.

    Science.gov (United States)

    Tseitlin, Mark; Yu, Zhelin; Quine, Richard W; Rinard, George A; Eaton, Sandra S; Eaton, Gareth R

    2014-12-01

    The use of multiple synchronized outputs from an arbitrary waveform generator (AWG) provides the opportunity to perform EPR experiments differently than by conventional EPR. We report a method for reconstructing the quadrature EPR spectrum from periodic signals that are generated with sinusoidal magnetic field modulation such as continuous wave (CW), multiharmonic, or rapid scan experiments. The signal is down-converted to an intermediate frequency (IF) that is less than the field scan or field modulation frequency and then digitized in a single channel. This method permits use of a high-pass analog filter before digitization to remove the strong non-EPR signal at the IF, that might otherwise overwhelm the digitizer. The IF is the difference between two synchronized X-band outputs from a Tektronix AWG 70002A, one of which is for excitation and the other is the reference for down-conversion. To permit signal averaging, timing was selected to give an exact integer number of full cycles for each frequency. In the experiments reported here the IF was 5kHz and the scan frequency was 40kHz. To produce sinusoidal rapid scans with a scan frequency eight times IF, a third synchronized output generated a square wave that was converted to a sine wave. The timing of the data acquisition with a Bruker SpecJet II was synchronized by an external clock signal from the AWG. The baseband quadrature signal in the frequency domain was reconstructed. This approach has the advantages that (i) the non-EPR response at the carrier frequency is eliminated, (ii) both real and imaginary EPR signals are reconstructed from a single physical channel to produce an ideal quadrature signal, and (iii) signal bandwidth does not increase relative to baseband detection. Spectra were obtained by deconvolution of the reconstructed signals for solid BDPA (1,3-bisdiphenylene-2-phenylallyl) in air, 0.2mM trityl OX63 in water, 15 N perdeuterated tempone, and a nitroxide with a 0.5G partially-resolved proton

  20. Prevalence and Severity of Off-Centering During Diagnostic CT: Observations From 57,621 CT scans of the Chest, Abdomen, and/or Pelvis.

    Science.gov (United States)

    Akin-Akintayo, Oladunni O; Alexander, Lauren F; Neill, Rebecca; Krupinksi, Elizabeth A; Tang, Xiangyang; Mittal, Pardeep K; Small, William C; Moreno, Courtney C

    2018-02-23

    To determine distances between patient centroid and gantry isocenter during CT imaging of the chest, abdomen, and/or pelvis, and to evaluate differences based on patient gender, scan region, patient position, and gantry aperture. A water phantom and an anthropomorphic phantom were imaged in the centered position in the CT gantry and at several off-centered positions. Additionally, data from 57,621 adult chest, abdomen, and/or pelvic CT acquisitions were evaluated. Data were analyzed with an analysis of variance using the centroid-to-isocenter data as the dependent variable and the other parameters as independent variables. The majority of patient acquisitions (83.7% (48271/57621)) were performed with the patient's centroid positioned below isocenter (mean 1.7 cm below isocenter (SD 1.8 cm); range 12.1 cm below to 7.8 cm above isocenter). Off-centering in the x-axis was less severe (mean 0.01 cm left of isocenter (SD 1.6 cm)). Distance between centroid and isocenter in the y-axis did not differ as a function of sex but did differ based on scan region, patient position, and gantry aperture. Off-centering is common during CT imaging and has been previously demonstrated to impact dose and image quality. Copyright © 2018 Elsevier Inc. All rights reserved.

  1. The value of digital tomosynthesis of the chest as a problem-solving tool for suspected pulmonary nodules and hilar lesions detected on chest radiography.

    Science.gov (United States)

    Galea, Angela; Dubbins, Paul; Riordan, Richard; Adlan, Tarig; Roobottom, Carl; Gay, David

    2015-05-01

    To assess the capability of digital tomosynthesis (DTS) of the chest compared to a postero-anterior (PA) and lateral chest radiograph (CXR) in the diagnosis of suspected but unconfirmed pulmonary nodules and hilar lesions detected on a CXR. Computed tomography (CT) was used as the reference standard. 78 patients with suspected non-calcified pulmonary nodules or hilar lesions on their CXR were included in the study. Two radiologists, blinded to the history and CT, prospectively analysed the CXR (PA and lateral) and the DTS images using a picture archiving and communication workstation and were asked to designate one of two outcomes: true intrapulmonary lesion or false intrapulmonary lesion. A CT of the chest performed within 4 weeks of the CXR was used as the reference standard. Inter-observer agreement and time to report the modalities were calculated for CXR and DTS. There were 34 true lesions confirmed on CT, 12 were hilar lesions and 22 were peripheral nodules. Of the 44 false lesions, 37 lesions were artefactual or due to composite shadow and 7 lesions were real but extrapulmonary simulating non-calcified intrapulmonary lesions. The PA and lateral CXR correctly classified 39/78 (50%) of the lesions, this improved to 75/78 (96%) with DTS. The sensitivity and specificity was 0.65 and 0.39 for CXR and 0.91 and 1 for DTS. Based on the DTS images, readers correctly classified all the false lesions but missed 3/34 true lesions. Two of the missed lesions were hilar in location and one was a peripheral nodule. All three missed lesions were incorrectly classified on DTS as composite shadow. DTS improves diagnostic confidence when compared to a repeat PA and lateral CXR in the diagnosis of both suspected hilar lesions and pulmonary nodules detected on CXR. DTS is able to exclude most peripheral pulmonary nodules but caution and further studies are needed to assess its ability to exclude hilar lesions. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  2. Slotted Circularly Polarized Microstrip Antenna for RFID Application

    Directory of Open Access Journals (Sweden)

    S. Kumar

    2017-12-01

    Full Text Available A single layer coaxial fed rectangular microstrip slotted antenna for circular polarization (CP is proposed for radio frequency identification (RFID application. Two triangular shaped slots and one rectangular slot along the diagonal axis of a square patch have been embedded. Due to slotted structure along the diagonal axis and less surface area, good quality of circular polarization has been obtained with the reduction in the size of microstrip antenna by 4.04 %. Circular polarization radiation performance has been studied by size and angle variation of diagonally slotted structures. The experimental result found for 10-dB return loss is 44 MHz with 10MHz of 3dB Axial Ratio (AR bandwidth respectively at the resonant frequency 910 MHz. The overall proposed antenna size including the ground plane is 80 mm x 80 mm x 4.572 mm.

  3. Reader performance in detection of pneumothorax on conventional chest films versus computed radiographs

    International Nuclear Information System (INIS)

    Elam, E.A.; Hillman, B.J.; Fajardo, L.L.

    1990-01-01

    This paper evaluates the suitability of digital radiography for the detection of pneumothoraces, we compared radiologist' performance in four interpretation settings: conventional film-screen (FS) chest radiographs; small-format (8 x 6.5-inch) computed radiography (CR) (Toshiba); large-format (14 x 17-inch) CR; and CR images viewed on a digital workstation. Twenty-three frontal-view chest radiographs with pneumothoraces and 22 other chest radiographs, either normal or showing miscellaneous abnormalities, were read by six board-certified radiologist in each interpretation setting. We compared the sensitivity, specificity, and receiver operating characteristic (ROC) performances they achieved among the interpretation settings with the use of sequential paired t tests. To date, four readers have completed their reading of both the FS and the small-format CR films

  4. Influence of Closed Stator Slots on Cogging Torque

    DEFF Research Database (Denmark)

    Ion, Trifu; Leban, Krisztina Monika; Ritchie, Ewen

    2013-01-01

    Cogging torque results due interaction of magnetic field of magnets and stator slots, and have negative effects on permanent magnet machines such as vibrations, noise, torque ripples and problems during turbine start-up and cut-in. In order to reduce cogging torque this paper presents a study...... of influence of closed stator slots on cogging torque using magnetic slot wedges....

  5. Comparison of Bolton analysis and Little’s irregularity index on laser scanned three-dimensional digital study models with conventional study models

    Science.gov (United States)

    Kurnia, H.; Noerhadi, N. A. I.

    2017-08-01

    Three-dimensional digital study models were introduced following advances in digital technology. This study was carried out to assess the reliability of digital study models scanned by a laser scanning device newly assembled. The aim of this study was to compare the digital study models and conventional models. Twelve sets of dental impressions were taken from patients with mild-to-moderate crowding. The impressions were taken twice, one with alginate and the other with polyvinylsiloxane. The alginate impressions were made into conventional models, and the polyvinylsiloxane impressions were scanned to produce digital models. The mesiodistal tooth width and Little’s irregularity index (LII) were measured manually with digital calipers on the conventional models and digitally on the digital study models. Bolton analysis was performed on each study models. Each method was carried out twice to check for intra-observer variability. The reproducibility (comparison of the methods) was assessed using independent-sample t-tests. The mesiodistal tooth width between conventional and digital models did not significantly differ (p > 0.05). Independent-sample t-tests did not identify statistically significant differences for Bolton analysis and LII (p = 0.603 for Bolton and p = 0894 for LII). The measurements of the digital study models are as accurate as those of the conventional models.

  6. Slot Antenna for Wireless Temperature Measurement Systems

    DEFF Research Database (Denmark)

    Acar, Öncel; Jakobsen, Kaj Bjarne

    2016-01-01

    This paper presents a novel clover-slot antenna for a surface-acoustic-wave sensor based wireless temperature measurement system. The slot is described by a parametric locus curve that has the shape of a clover. The antenna is operated at high temperatures, in rough environments, and has a 43......% fractional bandwidth at the 2.4 GHz ISM-band. The slot antenna has been optimized for excitation by a passive chip soldered onto it. Measurement results are compared with simulation results and show good agreements....

  7. Vertically Polarized Omnidirectional Printed Slot Loop Antenna

    DEFF Research Database (Denmark)

    Kammersgaard, Nikolaj Peter Iversen; Kvist, Søren H.; Thaysen, Jesper

    2015-01-01

    A novel vertically polarized omnidirectional printed slot loop antenna has been designed, simulated, fabricated and measured. The slot loop works as a magnetic loop. The loop is loaded with inductors to insure uniform and in-phase fields in the slot in order to obtain an omnidirectional radiation...... pattern. The antenna is designed for the 2.45 GHz Industrial, Scientific and Medical band. Applications of the antenna are many. One is for on-body applications since it is ideal for launching a creeping waves due to the polarization....

  8. Radiation dose reduction in the evaluation of scoliosis: an application of digital radiography

    International Nuclear Information System (INIS)

    Kushner, D.C.; Cleveland, R.H.; Herman, T.E.; Zaleske, D.J.; Ehrlich, M.G.; Correia, J.A.

    1986-01-01

    This report documents the clinical testing of scanning beam digital radiography as an imaging method in patients with scoliosis. This type of digital imaging requires a skin exposure of only 2.4 mR (0.619 microC/kg) per image, compared with the lowest possible posteroanterior screen-film exposure of 10 mR (2.58 microC/kg) at the chest and 60 mR (15.48 microC/kg) at the lumbar spine. Digital radiographic and screen-film images were obtained on multiple test objects and 273 patients. Scoliosis measurements using screen-film radiographs and digital radiographs were comparable to within a mean difference of 1 degrees at many different degrees of severity. The low-dose digital images were found to be useful and accurate for the detection and measurement of scoliosis after the first screen-film radiographs have excluded tumors and structural abnormalities

  9. G-tunnel pressurized slot-testing evaluations

    International Nuclear Information System (INIS)

    Zimmerman, R.M.; Sifre-Soto, C.; Mann, K.L.; Bellman, R.A. Jr.; Luker, S.; Dodds, D.J.

    1992-04-01

    Designers and analysts of radioactive waste repositories must be able to predict the mechanical behavior of the host rock. Sandia National Laboratories elected to conduct a development program to enhance mechanical-type measurements. The program was focused on pressurized slot testing and featured (1) development of an improved method to cut slots using a chain saw with diamond-tipped cutters, (2) measurements useful for determining in situ stresses normal to slots, (3) measurements applicable for determining the in situ modulus of deformation parallel to a drift surface, and (4) evaluations of pressurized slot strength testing results and methods. This report contains data interpretation and evaluations. Included are recommendations for future efforts. This third report contains the interpretations of the testing with emphasis on the measurement results as they apply to describing rock behavior. In particular, emphases are placed on (1) normal stress determinations using the flatjack cancellation (FC) method, (2) modulus of deformation determinations, and (3) high pressure investigations. Most of the material in the first two reports is not repeated here. Appropriate data are repeated in tabular form

  10. CT of blunt chest trauma in children

    International Nuclear Information System (INIS)

    Manson, D.; Babyn, P.S.; Palder, S.; Bergmann, K.

    1993-01-01

    While trauma is still the leading cause of death in the pediatric age range, it is surprising how little the CT appearances of pediatric chest injury have been investigated in the literature. We have reviewed the CT findings of blunt chest trauma in 44 children for whom chest CT examinations were requested to investigate the extent of intrathoracic injury. We noted a propensity for pulmonary contusions to be located posteriorly or posteromedially, and for them to be anatomically nonsegmental and crescentic in shape. This is possibly attributable to the relatively compliant anterior chest wall in children. The CT appearances of other major thoracic injuries are described, including pulmonary lacerations, pneumothoraces, malpositioned chest tubes, mediastinal hematomas, aortic injury, tracheobronchial injury, hemopericardium, and spinal injuries with paraspinal fluid collections. Children demonstrating findings incidental to the actual injury yet important to the subsequent therapy are also presented. We conclude that, in the event of clinically significant blunt chest trauma, the single supine chest examination in the trauma room is insufficient to adequately identify the extent of intrathoracic injury. With the exception of concern for aortic injury for which aortography is indicated, a dynamically enhanced CT scan of the thorax should be performed as clinically significant findings may result in altered therapy. (orig.)

  11. Computed tomography of chest trauma

    International Nuclear Information System (INIS)

    Dinkel, E.; Uhl, H.; Reinbold, W.D.; Wimmer, B.; Wenz, W.

    1987-01-01

    Chest CT scans were obtained in 86 patients suffering from serious blunt or penetrating chest trauma. The finding of mediastinal widening was by far the most common CT indication. CT proved to be a more sensitive method for detection of parenchymal lung lesions and occult pneumothorax than bedside radiographs. CT contributed substantially in differentiation of lung abscess and empyema, exclusion of mediastinal pathology and spinal injuries. Aortography is still indicated, even when CT findings are normal, if aortic laceration is clinically suspected. Despite all technical problems combined with CT examinations in the critically ill patient, we consider CT a valuable diagnostic tool for selected problems in the traumatized patient. (orig.) [de

  12. Computed tomography of chest trauma

    Energy Technology Data Exchange (ETDEWEB)

    Dinkel, E.; Uhl, H.; Reinbold, W.D.; Wimmer, B.; Wenz, W.

    1987-09-01

    Chest CT scans were obtained in 86 patients suffering from serious blunt or penetrating chest trauma. The finding of mediastinal widening was by far the most common CT indication. CT proved to be a more sensitive method for detection of parenchymal lung lesions and occult pneumothorax than bedside radiographs. CT contributed substantially in differentiation of lung abscess and empyema, exclusion of mediastinal pathology and spinal injuries. Aortography is still indicated, even when CT findings are normal, if aortic laceration is clinically suspected. Despite all technical problems combined with CT examinations in the critically ill patient, we consider CT a valuable diagnostic tool for selected problems in the traumatized patient.

  13. Nuclear imaging of the chest

    International Nuclear Information System (INIS)

    Bahk, Y.W.

    1998-01-01

    This book provides up-to-the minute information on the diagnostic nuclear imaging of chest disorders. The authors have endeavored to integrate and consolidate the many different subspecialities in order to enable a holistic understanding of chest diseases from the nuclear medicine standpoint. Highlights of the book include in addition to the cardiac scan the description of aerosol lung imaging in COPD and other important pulmonary diseases and the updates on breast and lung cancer imaging, as well as imaging of the bony thorax and esophagus. It is required reading not only for nuclear medicine practitioners and researchers but also for all interested radiologists, traumatologists, pulmonologists, oncologists and cardiologists. (orig.)

  14. Energy dissipation of slot-type flip buckets

    Science.gov (United States)

    Wu, Jian-hua; Li, Shu-fang; Ma, Fei

    2018-03-01

    The energy dissipation is a key index in the evaluation of energy dissipation elements. In the present work, a flip bucket with a slot, called the slot-type flip bucket, is theoretically and experimentally investigated by the method of estimating the energy dissipation. The theoretical analysis shows that, in order to have the energy dissipation, it is necessary to determine the sequent flow depth h 1 and the flow speed V 1 at the corresponding position through the flow depth h 2 after the hydraulic jump. The relative flow depth h 2 / h 。 is a function of the approach flow Froude number Fr 。, the relative slot width b/B 。, and the relative slot angle θ/β. The expression for estimating the energy dissipation is developed, and the maximum error is not larger than 9.21%.

  15. Chest wall tuberculosis; CT findings in 14 patients

    Energy Technology Data Exchange (ETDEWEB)

    Han, Young Min; Lee, Sang Yong; Lee, Dong Won; Kim, Chong Soo; Chung, Gyung Ho; Sohn, Myung Hee; Choi, Ki Chul [Chonbuk National Univ. School of Medicine, Chonju (Korea, Republic of); Kim, Dong Woo [Daejoen Eulji Hospital, Daejon (Korea, Republic of); Juhng, Seon Kwan [Wonkwang Univ. Medical School, Iksan (Korea, Republic of)

    1996-01-01

    To present CT findings of chest wall tuberculosis. CT scans were obtained in 14 patients with proven chest wall tuberculosis. Diagnosis was confirmed by means of right open thoracostomy with abscess evacuation (n=1), excision and curettage (n=11) or excision and curettage along with resection of the involved lung (n=2). The images were assessed with emphasis for the extrapleural, pleural, and pulmonary lesions. All patients showed juxtacostal soft tissue mass with central low attenuation and peripheral rim enhancement. The lesions were located in the left hemithorax in eight patients and in the right in six. Multiple lesions were found in three patients (two in one and three in two). Rib destruction was observed in four patients. Intercostal muscle involvement of thickening and enhancement were shown in all patients. Thirteen patients (93%) had evidence of pulmonary tuberculosis: active pulmonary tuberculosis in nine and stable tuberculosis in four. Pleural lesions, including empyema necessitatis in six, were observed in eleven (79%). On CT scan, chest wall tuberculosis is characterized by juxtacostal soft tissue lesion with central low attenuation and peripheral rim enhancement. Rib destruction may be associated. Additionally, enhancing intercostal muscle suggest direct inflammatory process of tuberculosis and spread channel to the chest wall involvement of pleuropulmonary tuberculosis.

  16. Chest wall tuberculosis; CT findings in 14 patients

    International Nuclear Information System (INIS)

    Han, Young Min; Lee, Sang Yong; Lee, Dong Won; Kim, Chong Soo; Chung, Gyung Ho; Sohn, Myung Hee; Choi, Ki Chul; Kim, Dong Woo; Juhng, Seon Kwan

    1996-01-01

    To present CT findings of chest wall tuberculosis. CT scans were obtained in 14 patients with proven chest wall tuberculosis. Diagnosis was confirmed by means of right open thoracostomy with abscess evacuation (n=1), excision and curettage (n=11) or excision and curettage along with resection of the involved lung (n=2). The images were assessed with emphasis for the extrapleural, pleural, and pulmonary lesions. All patients showed juxtacostal soft tissue mass with central low attenuation and peripheral rim enhancement. The lesions were located in the left hemithorax in eight patients and in the right in six. Multiple lesions were found in three patients (two in one and three in two). Rib destruction was observed in four patients. Intercostal muscle involvement of thickening and enhancement were shown in all patients. Thirteen patients (93%) had evidence of pulmonary tuberculosis: active pulmonary tuberculosis in nine and stable tuberculosis in four. Pleural lesions, including empyema necessitatis in six, were observed in eleven (79%). On CT scan, chest wall tuberculosis is characterized by juxtacostal soft tissue lesion with central low attenuation and peripheral rim enhancement. Rib destruction may be associated. Additionally, enhancing intercostal muscle suggest direct inflammatory process of tuberculosis and spread channel to the chest wall involvement of pleuropulmonary tuberculosis

  17. Evaluation of a Noise Reduction Procedure for Chest Radiography

    Science.gov (United States)

    Fukui, Ryohei; Ishii, Rie; Kodani, Kazuhiko; Kanasaki, Yoshiko; Suyama, Hisashi; Watanabe, Masanari; Nakamoto, Masaki; Fukuoka, Yasushi

    2013-01-01

    Background The aim of this study was to evaluate the usefulness of noise reduction procedure (NRP), a function in the new image processing for chest radiography. Methods A CXDI-50G Portable Digital Radiography System (Canon) was used for X-ray detection. Image noise was analyzed with a noise power spectrum (NPS) and a burger phantom was used for evaluation of density resolution. The usefulness of NRP was evaluated by chest phantom images and clinical chest radiography. We employed the Bureau of Radiological Health Method for scoring chest images while carrying out our observations. Results NPS through the use of NRP was improved compared with conventional image processing (CIP). The results in image quality showed high-density resolution through the use of NRP, so that chest radiography examination can be performed with a low dose of radiation. Scores were significantly higher than for CIP. Conclusion In this study, use of NRP led to a high evaluation in these so we are able to confirm the usefulness of NRP for clinical chest radiography. PMID:24574577

  18. OBLITERATION OF COSTOPHRENIC ANGTE IN A PLAIN X - RAY CHEST

    Directory of Open Access Journals (Sweden)

    Ramakrishna

    2015-08-01

    Full Text Available Obliteration of Costophrenicangle can be a pleural effusion in a majority of cases but other causes of mediastinal masses , aortic aneurysm , postpneumonectomy , lung and pleural masses and consolidation and collapse of the lung can cause shadows mimicking pleural effusion. It is always essential to take the help of later al and decubitus films , ultrasonography of chest and CT scan chest to come to a proper diagnosis. Inadvertent pleural aspiration basing on chest x - ray PA alone can have disastrous consequences

  19. 49 CFR 236.809 - Signal, slotted mechanical.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 4 2010-10-01 2010-10-01 false Signal, slotted mechanical. 236.809 Section 236.809 Transportation Other Regulations Relating to Transportation (Continued) FEDERAL RAILROAD... § 236.809 Signal, slotted mechanical. A mechanically operated signal with an electromagnetic device...

  20. A Waveguide Transverse Broad Wall Slot Radiating Between Baffles

    DEFF Research Database (Denmark)

    Dich, Mikael; Rengarajan, S.R.

    1997-01-01

    An analysis of the self impedance of waveguide-fed transverse slots radiating between baffles is presented. The region exterior to the slot is treated as a parallel plate (PP) waveguide which radiates into half space through an aperture in an infinite ground plane. The slot problem is analyzed...

  1. Terrestrial scanning or digital images in inventory of monumental objects? - case study

    Science.gov (United States)

    Markiewicz, J. S.; Zawieska, D.

    2014-06-01

    Cultural heritage is the evidence of the past; monumental objects create the important part of the cultural heritage. Selection of a method to be applied depends on many factors, which include: the objectives of inventory, the object's volume, sumptuousness of architectural design, accessibility to the object, required terms and accuracy of works. The paper presents research and experimental works, which have been performed in the course of development of architectural documentation of elements of the external facades and interiors of the Wilanów Palace Museum in Warszawa. Point clouds, acquired from terrestrial laser scanning (Z+F 5003h) and digital images taken with Nikon D3X and Hasselblad H4D cameras were used. Advantages and disadvantages of utilisation of these technologies of measurements have been analysed with consideration of the influence of the structure and reflectance of investigated monumental surfaces on the quality of generation of photogrammetric products. The geometric quality of surfaces obtained from terrestrial laser scanning data and from point clouds resulting from digital images, have been compared.

  2. Accuracy and reliability of a novel method for fusion of digital dental casts and cone beam computed tomography scans

    NARCIS (Netherlands)

    Rangel, F.A.; Maal, T.J.J.; Bronkhorst, E.M.; Breuning, K.H.; Schols, J.G.J.H.; Berge, S.J.; Kuijpers-Jagtman, A.M.

    2013-01-01

    Several methods have been proposed to integrate digital models into Cone Beam Computed Tomography scans. Since all these methods have some drawbacks such as radiation exposure, soft tissue deformation and time-consuming digital handling processes, we propose a new method to integrate digital dental

  3. Rectangular-cladding silicon slot waveguide with improved nonlinear performance

    Science.gov (United States)

    Huang, Zengzhi; Huang, Qingzhong; Wang, Yi; Xia, Jinsong

    2018-04-01

    Silicon slot waveguides have great potential in hybrid silicon integration to realize nonlinear optical applications. We propose a rectangular-cladding hybrid silicon slot waveguide. Simulation result shows that, with a rectangular-cladding, the slot waveguide can be formed by narrower silicon strips, so the two-photon absorption (TPA) loss in silicon is decreased. When the cladding material is a nonlinear polymer, the calculated TPA figure of merit (FOMTPA) is 4.4, close to the value of bulk nonlinear polymer of 5.0. This value confirms the good nonlinear performance of rectangular-cladding silicon slot waveguides.

  4. Imaging of chest disease due to intravenous heroin abuse

    International Nuclear Information System (INIS)

    Lian Xuhui; Chen Zhong; Ye Wenqin

    2002-01-01

    Objective: To study the imaging findings of the chest disease due to intravenous heroin abuse. Methods: Twenty-five cases of clinically confirmed chest disease due to intravenous heroin abuse were retrospectively analyzed. 25 cases had conventional X-ray film, 6 cases had CT scanning, and 6 cases had echocardiography scanning. Results: On X-ray and CT, the following signs were found: lung making manifold (n = 5), small patchy shadow (n = 15), pneumatocele (n = 16), small cavity (n = 16), small node (n = 7), pleural effusion (n = 8 ), pneumothorax (n = 2), hydropneumothorax (n = 6), pulmonary edema (n = 2), megacardia (n = 11), multiple-shaped lesion (n = 20). On echocardiography, tricuspid vegetation (n = 4) and tricuspid insufficiency (n = 4) were found. Conclusion: The X-ray and CT manifestations of chest inflammation due to intravenous heroin abuse are multiple. The multiple small cavities and pneumatoceles sign are of some value in the diagnosis of lung inflammation due to intravenous heroin abuse among young patients

  5. New Methods for Imaging Evaluation of Chest Wall Deformities

    Directory of Open Access Journals (Sweden)

    Ana Lain

    2017-12-01

    Full Text Available AimThe purpose of this study is to describe the development of an external 3-dimensional (3D scanner as a noninvasive method for imaging chest wall deformities. It allows objective assessment, reconstruction of the area of interest, and evaluation of the severity of the deformity by using external indexes.External 3D scanning systemThe OrtenBodyOne scanner (Orten, Lyon, France uses depth sensors to scan the entire 3D external body surface of a patient. The depth sensors combine structured light with two classic computer vision techniques: depth from focus and depth from stereo. The data acquired are processed and analyzed using the Orten-Clinic software.Materials and methodsTo investigate the performance of the device, a preliminary prospective study (January 2015–March 2016 was carried out in patients attending our hospital chest wall deformities unit. In total, 100 patients (children and young adults with pectus excavatum or pectus carinatum, treated by surgery or non-operative methods were included. In patients undergoing non-operative treatment, external 3D scanning was performed monthly until complete correction was achieved. In surgically treated patients, scanning was done before and after surgical correction. In 42 patients, computed tomography (CT was additionally performed and correlations between the Haller index calculated by CT and the external Haller index using external scanning were investigated using a Student’s test (r = 0.83.ConclusionExternal scanning is an effective, objective, radiation-free means to diagnose and follow-up patients with chest wall deformities. Externally measured indexes can be used to evaluate the severity of these conditions and the treatment outcomes.

  6. Slot Optimization Design of Induction Motor for Electric Vehicle

    Science.gov (United States)

    Shen, Yiming; Zhu, Changqing; Wang, Xiuhe

    2018-01-01

    Slot design of induction motor has a great influence on its performance. The RMxprt module based on magnetic circuit method can be used to analyze the influence of rotor slot type on motor characteristics and optimize slot parameters. In this paper, the authors take an induction motor of electric vehicle for a typical example. The first step of the design is to optimize the rotor slot by RMxprt, and then compare the main performance of the motor before and after the optimization through Ansoft Maxwell 2D. After that, the combination of optimum slot type and the optimum parameters are obtained. The results show that the power factor and the starting torque of the optimized motor have been improved significantly. Furthermore, the electric vehicle works at a better running status after the optimization.

  7. The reality of virtual anthropology: Comparing digitizer and laser scan data collection methods for the quantitative assessment of the cranium.

    Science.gov (United States)

    Algee-Hewitt, Bridget F B; Wheat, Amber D

    2016-05-01

    The use of geometric morphometry to study cranial variation has steadily grown in appeal over the past decade in biological anthropology. Publication trends suggest that the most popular methods for three-dimensional data acquisition involve landmark-based coordinate data collection using a digitizer. Newer laser scan approaches are seeing increasing use, owing to the benefits that densely sampled data offer. While both of these methods have their utility, research that investigates their compatibility is lacking. The purpose of this project is to compare, quantitatively, craniometrics collected with a digitizer against data extracted from laser scans using the same individuals and laboratory conditions. Three-dimensional (x,y,z) coordinates and traditional inter-landmark distances (ILDs) were obtained with a Microscribe digitizer and 360° color models produced from NextEngine laser scans for 38 adult crania representing five cemeteries from the ADBOU skeletal collection in Denmark. Variance-based tests were performed to evaluate the disagreement between data collected with a digitizer and from laser scan models. Consideration was given to differences among landmarks by type, between ILDs calculated from landmark coordinates, and in morphology for the cemetery populations. Further, the reliability of laser scan data collection was assessed by intra-observer error tests. Researchers should be aware of the potential error associated with the use of Types II and III landmarks and the limitations on reliability imposed by object-to-scanner placement. This project reveals how laser scans can provide a valuable digital archive of cranial material that can be reasonably exploited for the "virtual" collection of coordinates and the calculation of ILDs. © 2015 Wiley Periodicals, Inc.

  8. Chest radiography: new technological developments and their applications

    NARCIS (Netherlands)

    Schalekamp, S.; Ginneken, B. van; Karssemeijer, N.; Schaefer-Prokop, C.M.

    2014-01-01

    Digital chest radiography is still the most common radiological examination. With the upcoming three-dimensional (3D) acquisition techniques the value of radiography seems to diminish. But because radiography is inexpensive, readily available, and requires very little dose, it is still being used

  9. Predictors of return rate discrimination in slot machine play.

    Science.gov (United States)

    Coates, Ewan; Blaszczynski, Alex

    2014-09-01

    The purpose of this study was to investigate the extent to which accurate estimates of payback percentages and volatility combined with prior learning, enabled players to successfully discriminate between multi-line/multi-credit slot machines that provided differing rates of reinforcement. The aim was to determine if the capacity to discriminate structural characteristics of gaming machines influenced player choices in selecting 'favourite' slot machines. Slot machine gambling history, gambling beliefs and knowledge, impulsivity, illusions of control, and problem solving style were assessed in a sample of 48 first year undergraduate psychology students. Participants were subsequently exposed to a choice paradigm where they could freely select to play either of two concurrently presented PC-simulated slot machines programmed to randomly differ in expected player return rates (payback percentage) and win frequency (volatility). Results suggest that prior learning and cognitions (particularly gambler's fallacy) but not payback, were major contributors to the ability of a player to discriminate volatility between slot machines. Participants displayed a general tendency to discriminate payback, but counter-intuitively placed more bets on the slot machine with lower payback percentage rates.

  10. The Role of Auditory Features Within Slot-Themed Social Casino Games and Online Slot Machine Games.

    Science.gov (United States)

    Bramley, Stephanie; Gainsbury, Sally M

    2015-12-01

    Over the last few years playing social casino games has become a popular entertainment activity. Social casino games are offered via social media platforms and mobile apps and resemble gambling activities. However, social casino games are not classified as gambling as they can be played for free, outcomes may not be determined by chance, and players receive no monetary payouts. Social casino games appear to be somewhat similar to online gambling activities in terms of their visual and auditory features, but to date little research has investigated the cross over between these games. This study examines the auditory features of slot-themed social casino games and online slot machine games using a case study design. An example of each game type was played on three separate occasions during which, the auditory features (i.e., music, speech, sound effects, and the absence of sound) within the games were logged. The online slot-themed game was played in demo mode. This is the first study to provide a qualitative account of the role of auditory features within a slot-themed social casino game and an online slot machine game. Our results found many similarities between how sound is utilised within the two games. Therefore the sounds within these games may serve functions including: setting the scene for gaming, creating an image, demarcating space, interacting with visual features, prompting players to act, communicating achievements to players, providing reinforcement, heightening player emotions and the gaming experience. As a result this may reduce the ability of players to make a clear distinction between these two activities, which may facilitate migration between games.

  11. [Influence of slot size on torque control].

    Science.gov (United States)

    Tian, Jun; Liu, Zhong-Hao; Zhang, Ding; Wu, Chuan-Jun

    2009-12-01

    To study the influence of two slot size brackets on torque control when teeth interacted in the same arch. After the upper arch was aligned and leveled in Typodont study, the inclinations of upper teeth 5 +/- 5 were measured when 0.457 2 mm x 0.635 0 mm OPA-K brackets and 0.558 8 mmx0.711 2 mm OPA-K brackets were filled with 0.431 8 mm x 0.635 0 mm stainless steel wire. This experiment was duplicated 10 times. The inclin of each tooth were transformed to the absolute values of the torque play angle psi by computing program, and paired-t test was used. The two kinds of slot size brackets were different with statistical significance on torque control. When the brackets were filled with 0.431 8 mm x 0.635 0 mm stainless steel wire, the absolute values of the angle psi in 0.558 8 mm x 0.711 2 mm and 0.457 2 mm x 0.635 0 mm slot size brackets were 6.140 degrees +/- 3.758 degrees and 2.608 degrees +/- 1.479 degrees respectively, and the average difference of that between the two slot size brackets was 3.532 degrees. The absolute values of the angle psi in the upper left and right canine brackets were 2.560 degrees +/- 2.605 degrees, 4.230 degrees +/- 2.817 degrees, 1.260 degrees +/- 0.747 degrees and 2.070 degrees +/- 0.663 degrees respectively, and average differences between them were smaller than that in the other teeth. There was difference between the two kinds of slot size brackets on torque control, and 0.457 2 mm x 0.635 0 mm slot size bracket controls torque better when filled with the same size wire. In this study, the teeth interaction in the same arch probably caused the result that the difference of two slot size brackets on torque control was less than the study results of the theory calculations and material studys before.

  12. Efficacy study of the digital image processing with varying pixel size at A/D conversion of the chest radiograph

    Energy Technology Data Exchange (ETDEWEB)

    Inamoto, Kazuo; Tanaka, Shinichi; Miura, Takashi; Takahashi, Akira; Iwata, Tetsuya

    1984-12-01

    In the study for development of medical image archiving system, we made experiments in the field of conversion of a X-ray picture to a digital form. Three sets of chest radiograph were selected for the study of digitalization by reading different pixel size (150-300 m) in the method of A/D conversion using a drum scanner and again reconstructed to an analog form after D/A conversion. These copy films of different pixel size were shown and evaluated by 48 volunteer doctors to choose a favorite picture. It did not always follow that the most favorite picture was the finest one using the smallest pixel size. This discrepancy was further analyzed by measurement of a density histogram. By comparison studies of density curves in the same ROI of different pixel size pictures, it was concluded that their choices were dependent on not only fineness but also contrast of an output image after D/A conversion. Often better contrast picture was a key to the selection more than the pixel size. This indicates that digital storing radiographs will be possibly regenerated to the real image by the skillful operation of pixel size and contrast of a radiograph. The results will contribute to the process of recording analog X-ray picture by a digital form in the medical image archiving system.

  13. Technique for evaluation of spatial resolution and microcalcifications in digital and scanned images of a standard breast phantom

    International Nuclear Information System (INIS)

    Santana, Priscila do C.; Gomes, Danielle S.; Oliveira, Marcio A.; Oliveira, Paulo Marcio C. de; Meira-Belo, Luiz C.; Nogueira-Tavares, Maria S.

    2011-01-01

    In this work, an automated methodology to evaluate digital and scanned images of a standard phantom (Phantom Mama) was studied. The Phantom Mama was used as an important tool to check the quality of mammographs. The scanned images were digitized using a ScanMaker 9800XL, with resolution of 900 dpi. The aim of this work is to test an automatic methodology for evaluation of spatial resolution and microcalcifications group of phantom mama images acquired with the same parameters in the same equipment. In order to analyze the images we have used the ImageJ software (in Java) which is public domain. We have used the Fast Fourier transform technique to evaluate the spatial resolution and used the ImageJ function Subtract Background and the Light Background plus Sliding Paraboloid on the evaluation of the five groups of microcalcifications on the breast phantom to assess the viability of using automated methods for both types of images. The methodology was adequate for evaluated the microcalcifications group and the spatial resolution in scanned and digital images, but the Phantom Mama doesn't provide sufficient parameters to evaluate the spatial resolution in this images. (author)

  14. Reassessing the standard chest radiograph for intraparenchymal activity

    International Nuclear Information System (INIS)

    Putman, C.E.; Hoeck, B.

    1986-01-01

    Despite the recent enthusiasm for better means of determining the extent of the active inflammatory response in the lung parenchyma of sarcoidosis patients, standard chest radiography remains a reliable method of determining staging, and parenchymal activity as well. The chest radiograph is characteristically used to determine the severity and course of sarcoidosis, but most studies imply there are poor correlations between the radiographic opacities and other indicators such as physiological parameters, gallium scans, and computed tomography and bronchoalveolar lavage results. In this preliminary communication, the authors report a new method of determining sarcoidosis activity, as compared to gallium scan, based solely on the standard chest radiograph. This method is based on the presence or absence of air bronchogram abnormalities, peribronchial cuffing, and subpleural thickening. Adenopathy and linear, reticulonodular, and nodular opacities are not included in this system. In 85 recent studies, which were performed in 51 patients with biopsy-proven sarcoidosis, this staging method agreed with gallium scanning in 68.2% of the patients. This system failed in specific and identifiable subgroups of patients that we will delineate. The authors will not report the comparative indices with a pulmonary function analysis or bronchoalveolar lavage results. Their database has not been tabulated to conclusively present these results, but it is their belief that this methodology will probably have a better correlation with lavage results and small airway function than would results of gallium scans. Of singular importance is the fact that this method can, with a more than acceptable degree of accuracy, distinguish between ''low-intensity'' alveolitis and ''high-intensity'' alveolitis

  15. Digital Processing for Modifying and Rearranging Rectilinear and Section Scan Data under Direct Observation

    Energy Technology Data Exchange (ETDEWEB)

    Kuhl, D. E.; Edwards, R. Q. [University of Pennsylvania, Philadelphia, PA (United States)

    1969-01-15

    Our digital processor for scan data is an on-site instrument that is intermediate in complexity between conventional optical processing devices and large digital computers. It is designed to provide for a wide and flexible range of secondary data operations, direct picture display on a CRT screen, and full operator control of both processing and display operations at the time of viewing. The instrument does not require the user to learn complicated programing schemes. The operator is expected to be a physician who will control the parameters of interest by punching preset buttons on a keyboard while observing changes displayed on a CRT screen. The system functions primarily as an investigative tool for studying perception of scan information and ways of making this information more meaningful. Data operations include data bounding, spatial averaging, iso-count line generation, image addition and subtraction, and several forms of quantitative read-out for analysis of regional data. The instrument is intended to serve as a central processor and reader for data from several units. Investigations with this processor have served as a source of information leading to the design of more simple processing devices suitable for wider acceptance. For example, the Mark III rectilinear and transverse section brain scanner that has evolved from this project is expected to be a practical improvement of the brain study method. This instrument is designed especially for rapid brain scanning using {sup 99m}Tc pertechnetate. It has a self-contained computer, integrated digital circuits for compactness and economy, and provision for transverse section scanning. The advantages of this system are that it provides a more thorough study using both transverse section and rectilinear modes, rapid performance, precise orientation of section and rectilinear views to the patient position, efficient transfer of information between physician and machine during studies, and economy of design

  16. Inferior hilar window on lateral chest radiographs

    International Nuclear Information System (INIS)

    Park, C.K.; Webb, W.R.; Klein, J.S.

    1990-01-01

    This paper determines the accuracy of lateral chest radiography in the detection of masses in the inferior hilar window, a normally avascular hilar region anterior to the lower lobe bronchi. Fifty patients with normal thoracic CT scans and 25 with hilar masses/adenopathy were selected retrospectively. The 75 corresponding lateral chest radiographs were blindly evaluated for visibility of the anterior walls of the lower lobe bronchi and the presence and laterality of abnormal soft tissue (>1 cm) in the inferior hilar window. Only a 7 x 7-cm square of the lateral radiograph was viewed

  17. The Gambling Reducing Slot Machine

    DEFF Research Database (Denmark)

    Callesen, Mette Buhl; Thomsen, Kristine Rømer; Linnet, Jakob

    2007-01-01

      The Gambling Reducing Slot Machine - Preliminary results Mette Buhl Callesen, Kristine Rømer Thomsen, Jakob Linnet and Arne Møller The PET Centre, Aarhus University Hospital and Centre of Functionally Integrative Neuroscience, Aarhus, Denmark   Slot machines are among the most addictive forms...... and willingness to continue gambling. The results may have important implications for understanding how to reduce gambling behavior in pathological gamblers.   [1] Griffiths, M. 1999. Gambling Technologies: Prospects for Problem Gambling. Journal of Gambling Studies, vol. 15(3), pp. 265-283.    ...

  18. Evaluation of digital dental models obtained from dental cone-beam computed tomography scan of alginate impressions.

    Science.gov (United States)

    Jiang, Tingting; Lee, Sang-Mi; Hou, Yanan; Chang, Xin; Hwang, Hyeon-Shik

    2016-05-01

    To investigate the dimensional accuracy of digital dental models obtained from the dental cone-beam computed tomography (CBCT) scan of alginate impressions according to the time elapse when the impressions are stored under ambient conditions. Alginate impressions were obtained from 20 adults using 3 different alginate materials, 2 traditional alginate materials (Alginoplast and Cavex Impressional) and 1 extended-pour alginate material (Cavex ColorChange). The impressions were stored under ambient conditions, and scanned by CBCT immediately after the impressions were taken, and then at 1 hour intervals for 6 hours. After reconstructing three-dimensional digital dental models, the models were measured and the data were analyzed to determine dimensional changes according to the elapsed time. The changes within the measurement error were regarded as clinically acceptable in this study. All measurements showed a decreasing tendency with an increase in the elapsed time after the impressions. Although the extended-pour alginate exhibited a less decreasing tendency than the other 2 materials, there were no statistically significant differences between the materials. Changes above the measurement error occurred between the time points of 3 and 4 hours after the impressions. The results of this study indicate that digital dental models can be obtained simply from a CBCT scan of alginate impressions without sending them to a remote laboratory. However, when the impressions are not stored under special conditions, they should be scanned immediately, or at least within 2 to 3 hours after the impressions are taken.

  19. Digitization of medical documents: an X-Windows application for fast scanning.

    Science.gov (United States)

    Muñoz, A; Salvador, C H; Gonzalez, M A; Dueñas, A

    1992-01-01

    This paper deals with digitization, using a commercial scanner, of medical documents as still images for introduction into a computer-based Information System. Document management involves storing, editing and transmission. This task has usually been approached from the perspective of the difficulties posed by radiologic images because of their indisputable qualitative and quantitative significance. However, healthcare activities require the management of many other types of documents and involve the requirements of numerous users. One key to document management will be the availability of a digitizer to deal with the greatest possible number of different types of documents. This paper describes the relevant aspects of documents and the technical specifications that digitizers must fulfill. The concept of document type is introduced as the ideal set of digitizing parameters for a given document. The use of document type parameters can drastically reduce the time the user spends in scanning sessions. Presentation is made of an application based on Unix, X-Windows and OSF/Motif, with a GPIB interface, implemented around the document type concept. Finally, the results of the evaluation of the application are presented, focusing on the user interface, as well as on the viewing of color images in an X-Windows environment and the use of lossy algorithms in the compression of medical images.

  20. An improved three-dimensional non-scanning laser imaging system based on digital micromirror device

    Science.gov (United States)

    Xia, Wenze; Han, Shaokun; Lei, Jieyu; Zhai, Yu; Timofeev, Alexander N.

    2018-01-01

    Nowadays, there are two main methods to realize three-dimensional non-scanning laser imaging detection, which are detection method based on APD and detection method based on Streak Tube. However, the detection method based on APD possesses some disadvantages, such as small number of pixels, big pixel interval and complex supporting circuit. The detection method based on Streak Tube possesses some disadvantages, such as big volume, bad reliability and high cost. In order to resolve the above questions, this paper proposes an improved three-dimensional non-scanning laser imaging system based on Digital Micromirror Device. In this imaging system, accurate control of laser beams and compact design of imaging structure are realized by several quarter-wave plates and a polarizing beam splitter. The remapping fiber optics is used to sample the image plane of receiving optical lens, and transform the image into line light resource, which can realize the non-scanning imaging principle. The Digital Micromirror Device is used to convert laser pulses from temporal domain to spatial domain. The CCD with strong sensitivity is used to detect the final reflected laser pulses. In this paper, we also use an algorithm which is used to simulate this improved laser imaging system. In the last, the simulated imaging experiment demonstrates that this improved laser imaging system can realize three-dimensional non-scanning laser imaging detection.

  1. Pleuritic Chest Pain: Sorting Through the Differential Diagnosis.

    Science.gov (United States)

    Reamy, Brian V; Williams, Pamela M; Odom, Michael Ryan

    2017-09-01

    Pleuritic chest pain is characterized by sudden and intense sharp, stabbing, or burning pain in the chest when inhaling and exhaling. Pulmonary embolism is the most common serious cause, found in 5% to 21% of patients who present to an emergency department with pleuritic chest pain. A validated clinical decision rule for pulmonary embolism should be employed to guide the use of additional tests such as d-dimer assays, ventilation-perfusion scans, or computed tomography angiography. Myocardial infarction, pericarditis, aortic dissection, pneumonia, and pneumothorax are other serious causes that should be ruled out using history and physical examination, electrocardiography, troponin assays, and chest radiography before another diagnosis is made. Validated clinical decision rules are available to help exclude coronary artery disease. Viruses are common causative agents of pleuritic chest pain. Coxsackieviruses, respiratory syncytial virus, influenza, parainfluenza, mumps, adenovirus, cytomegalovirus, and Epstein-Barr virus are likely pathogens. Treatment is guided by the underlying diagnosis. Nonsteroidal anti-inflammatory drugs are appropriate for pain management in those with virally triggered or nonspecific pleuritic chest pain. In patients with persistent symptoms, persons who smoke, and those older than 50 years with pneumonia, it is important to document radiographic resolution with repeat chest radiography six weeks after initial treatment.

  2. Theory of a Traveling Wave Feed for a Planar Slot Array Antenna

    Science.gov (United States)

    Rengarajan, Sembiam

    2012-01-01

    Planar arrays of waveguide-fed slots have been employed in many radar and remote sensing applications. Such arrays are designed in the standing wave configuration because of high efficiency. Traveling wave arrays can produce greater bandwidth at the expense of efficiency due to power loss in the load or loads. Traveling wave planar slot arrays may be designed with a long feed waveguide consisting of centered-inclined coupling slots. The feed waveguide is terminated in a matched load, and the element spacing in the feed waveguide is chosen to produce a beam squinted from the broadside. The traveling wave planar slot array consists of a long feed waveguide containing resonant-centered inclined coupling slots in the broad wall, coupling power into an array of stacked radiating waveguides orthogonal to it. The radiating waveguides consist of longitudinal offset radiating slots in a standing wave configuration. For the traveling wave feed of a planar slot array, one has to design the tilt angle and length of each coupling slot such that the amplitude and phase of excitation of each radiating waveguide are close to the desired values. The coupling slot spacing is chosen for an appropriate beam squint. Scattering matrix parameters of resonant coupling slots are used in the design process to produce appropriate excitations of radiating waveguides with constraints placed only on amplitudes. Since the radiating slots in each radiating waveguide are designed to produce a certain total admittance, the scattering (S) matrix of each coupling slot is reduced to a 2x2 matrix. Elements of each 2x2 S-matrix and the amount of coupling into the corresponding radiating waveguide are expressed in terms of the element S11. S matrices are converted into transmission (T) matrices, and the T matrices are multiplied to cascade the coupling slots and waveguide sections, starting from the load end and proceeding towards the source. While the use of non-resonant coupling slots may provide an

  3. Radiation from a Slot System in the Coaxial Line Shield

    Science.gov (United States)

    Katrich, V. A.; Lyashchenko, V. A.; Medvedev, N. V.

    2012-06-01

    The problem of electromagnetic wave excitation, scattering and radiation by the system of transverse slots, cut in the outer conductor of an infinite coaxial line, is solved by the magnetomotive forces method. The radiation and reflection coefficients of the circular and arc slot systems are investigated in dependence on slot sizes and feeder parameters. The processes of radiation into lossy material media are studied. The researches have been carried out with the interconnection between slots of internal and external regions considered.

  4. Cardiac CT for the assessment of chest pain: Imaging techniques and clinical results

    International Nuclear Information System (INIS)

    Becker, Hans-Christoph; Johnson, Thorsten

    2012-01-01

    Immediate and efficient risk stratification and management of patients with acute chest pain in the emergency department is challenging. Traditional management of these patients includes serial ECG, laboratory tests and further on radionuclide perfusion imaging or ECG treadmill testing. Due to the advances of multi-detector CT technology, dedicated coronary CT angiography provides the potential to rapidly and reliably diagnose or exclude acute coronary artery disease. Life-threatening causes of chest pain, such as aortic dissection and pulmonary embolism can simultaneously be assessed with a single scan, sometimes referred to as “triple rule out” scan. With appropriate patient selection, cardiac CT can accurately diagnose heart disease or other sources of chest pain, markedly decrease health care costs, and reliably predict clinical outcomes. This article reviews imaging techniques and clinical results for CT been used to evaluate patients with chest pain entering the emergency department.

  5. Nodule detection in digital chest radiography: Effect of anatomical noise

    International Nuclear Information System (INIS)

    Baath, M.; Haakansson, M.; Boerjesson, S.; Hoeschen, C.; Tischenko, O.; Kheddache, S.; Vikgren, J.; Maansson, L. G.

    2005-01-01

    The image background resulting from imaged anatomy can be divided into those components that are meaningful to the observers, in the sense that they are recognised as separate structures, and those that are not. These latter components (referred to as anatomical noise) can be removed using a method developed within the RADIUS group. The aim of the present study was to investigate whether the removal of the anatomical noise results in images where lung nodules with lower contrast can be detected. A receiver operating characteristic (ROC) study was therefore conducted using two types of images: clinical chest images and chest images in which the anatomical noise had been removed. Simulated designer nodules with a full-width-at-fifth-maximum of 10 mm but with varying contrast were added to the images. The contrast needed to obtain an area under the ROC curve of 0.80, C0.8, was used as a measure of detectability (a low value of C0.8 represents a high delectability). Five regions of the chest X ray were investigated and it was found that in all regions the removal of anatomical noise led to images with lower C0.8 than the original images. On average, C0.8 was 20% higher in the original images, ranging from 7% (the lateral pulmonary regions) to 41% (the upper mediastinal regions). (authors)

  6. Doses from pediatric CT examinations in Norway: are pediatric scan protocols developed and in daily use?

    International Nuclear Information System (INIS)

    Friberg, Eva G.

    2008-01-01

    Doses to pediatric patients from CT examinations are known to be unnecessarily high if scan protocols developed for adult patients are adopted. This overexposure is most often not recognized by the operating radiographer, due to the digital behavior of the imaging system. Use of optimized size-specific pediatric scan protocols is therefore essential to keep the doses at an appropriate level. The aim of this study was to investigate the doses to pediatric patients from CT examinations and to evaluate the level of optimization of the scan protocols. Patient data, applied scan parameters together with the dose parameters volume computed tomography dose index (CTD vol ) and dose length product (DLP) for examinations of the head, chest and abdomen were collected by means of a questionnaire from five university hospitals. The effective dose was estimated from the total DLP by use of region-specific conversion coefficients (E DLP ). Totally 136, 108 and 82 questionnaires were received for examinations of the head, chest and abdomen, respectively. Large variations in patient doses between the hospitals were observed, addressing the need for optimization of the scan protocols in general. Most of the hospitals applied successive lower mAs with decreasing patient age for all scan areas, while the use of lower tube voltage for small patients and a higher tube voltage for large patients were more rarely. This indicates the presence, to a certain level, of size specific scan protocols at some Norwegian hospitals. Focus on developing size-specific scan protocols for pediatric patients are important to reduce the doses and risks associated with pediatric CT examinations. (author)

  7. Multi-detector row computed tomography and blunt chest trauma

    International Nuclear Information System (INIS)

    Scaglione, Mariano; Pinto, Antonio; Pedrosa, Ivan; Sparano, Amelia; Romano, Luigia

    2008-01-01

    Blunt chest trauma is a significant source of morbidity and mortality in industrialized countries. The clinical presentation of trauma patients varies widely from one individual to another and ranges from minor reports of pain to shock. Knowledge of the mechanism of injury, the time of injury, estimates of motor vehicle accident velocity and deceleration, and evidence of associated injury to other systems are all salient features to provide for an adequate assessment of chest trauma. Multi-detector row computed tomography (MDCT) scanning and MDCT-angiography are being used more frequently in the diagnosis of patients with chest trauma. The high sensitivity of MDCT has increased the recognized spectrum of injuries. This new technology can be regarded as an extremely valuable adjunct to physical examination to recognize suspected and unsuspected blunt chest trauma

  8. Image quality of a Konica Regius 336 digital system in chest radiography; Qualita' dell'immagine di un sistema digitale Konica Regius 336 per radiologia toracica

    Energy Technology Data Exchange (ETDEWEB)

    Ostinelli, A.; Frigerio, M.; Monti, A.F.; Gelosa, S.; Tognoli, P.; Perniola, N. [Azienda Ospedaliera S. Anna, Como (Italy). Servizio di Fisica Sanitaria; Gozzi, G. [Azienda Ospedaliera S. Anna, Como (Italy). Servizio di Radiologia

    2000-06-01

    Digital radiographic systems permit to optimize execution, depiction and storage of radiological images. Since a Regius 336 digital system (Konica Corp. Tokyo, Japan) devoted to chest radiography Department of S. Anna Hospital in Como, Italy, it was investigated its performance relative to image quality. Konica Regius 336 is a computed radiography system made of a phosphorescence detector plate which is scanned with an infrared semiconductor laser beam. The radiographic image obtained from the detector is subjected to image processing, which allows a stable output and the nonlinear curve typical of conventional radiographic systems. Image quality was assessed based on the following parameters: dose, contrast, noise and spatial resolution. As reference, it was assessed the same parameters on a Cronex 88 analogic chest-changer (DuPont Pharma, North Billerica, Mass, USA). The Regius 336 air kerma values were always higher than the analogic ones (about 10%), both with and without a chest phantom; noise was also greater than in analogic images, sometimes even doubled. The optical densities of a step wedge and the spatial resolution of the digital chest-changer are independent of the X-ray tube voltage consequent to broader optical latitude. Inversely, the analogic images of the wedges show great optical density variability as a function of the X-ray tube voltage (in a range of 2). The modulation transfer functions of the two systems have the same trend. The performance of the Konica Regius 336 is nearly equivalent to that of an analogic system. The main advantages of the digital system are a standard output, lower consumption of radiographic films, higher productiveness and better image quality standard level. [Italian] I sistemi radiografici digitali offrono la possibilita' di ottimizzare l'esecuzione, la visualizzazione e l'archiviazione dell'immagine radiologica. Poiche' presso il Servizio di Radiologia dell'Azienda Ospedaliera S. Anna

  9. Investigation of a slot nanoantenna in optical frequency range

    Science.gov (United States)

    Dinesh kumar, V.; Asakawa, Kiyoshi

    2009-11-01

    Following the analogy of radio frequency slot antenna and its complementary dipole, we propose the implementation of a slot nanoantenna (SNA) in the optical frequency range. Using finite-difference time-domain (FDTD) method, we investigate the electromagnetic (EM) properties of a SNA formed in a thin gold film and compare the results with the properties of a gold dipole nanoantenna (DNA) of the same dimension as the slot. It is found that the response of the SNA is very similar to the DNA, like their counterparts in the radio frequency (RF) range. The SNA can enhance the near field intensity of incident field which strongly depends on its feedgap dimension. The resonance of the SNA is influenced by its slot length; for the increasing slot length, resonant frequency decreases whereas the sharpness of resonance increases. Besides, the resonance of the SNA is found sensitive to the thickness of metal film, when the latter is smaller than the skin depth. The effect of polarization of incident field on the EM response of the SNA was examined; the field enhancement is optimum when polarization is parallel to the feedgap. Finally, we calculate the radiation patterns of the DNA and SNA and compare them with those of the RF dipole antenna. The radiation pattern of the SNA is found to be independent of its slot length when excited at resonant frequency. To the best of our knowledge, this is the first study on a slot antenna in the optical frequency.

  10. Metal membrane with dimer slots as a universal polarizer

    Science.gov (United States)

    Zhukovsky, Sergej; Zalkovskij, Maksim; Malureanu, Radu; Kremers, Christian; Chigrin, Dmitry; Tang, Peter T.; Jepsen, Peter U.; Lavrinenko, Andrei V.

    2014-03-01

    In this work, we show theoretically and confirm experimentally that thin metal membranes patterned with an array of slot dimers (or their Babinet analogue with metal rods) can function as a versatile spectral and polarization filter. We present a detailed covariant multipole theory for the electromagnetic response of an arbitrary dimer based on the Green functions approach. The theory confirms that a great variety of polarization properties, such as birefringence, chirality and elliptical dichroism, can be achieved in a metal layer with such slot-dimer patterning (i.e. in a metasurface). Optical properties of the metasurface can be extensively tuned by varying the geometry (shape and dimensions) of the dimer, for example, by adjusting the sizes and mutual placement of the slots (e.g. inter-slot distance and alignment angle). Three basic shapes of dimers are analyzed: II-shaped (parallel slots), V-shaped, and T-shaped. These particular shapes of dimers are found to be sensitive to variations of the slots lengths and orientation of elements. Theoretical results are well supported by full-wave three-dimensional simulations. Our findings were verified experimentally on the metal membranes fabricated using UV lithography with subsequent Ni growth. Such metasurfaces were characterized using time-domain THz spectroscopy. The samples exhibit pronounced optical activity (500 degrees per wavelength) and high transmission: even though the slots cover only 4.3 % of the total membrane area the amplitude transmission reaches 0.67 at the resonance frequency 0.56 THz.

  11. Military Interoperable Digital Hospital Testbed

    Science.gov (United States)

    2012-07-01

    Subject outcome measures include blood pressure, waist circumference, weight, body mass index (BMI), body fat, HDL cholesterol, triglycerides , glucose ...master patient index (MPI), 625 duplicate chest x-rays and CT scans of the head between sending and receiving institution (taken within 0-7 days) were... Index (MPI) software. The report included chest x-rays (CPT 71010 and 71020) and CT scans of the head (CPT 70450) for the stated time periods. The

  12. Humanizing machines: Anthropomorphization of slot machines increases gambling.

    Science.gov (United States)

    Riva, Paolo; Sacchi, Simona; Brambilla, Marco

    2015-12-01

    Do people gamble more on slot machines if they think that they are playing against humanlike minds rather than mathematical algorithms? Research has shown that people have a strong cognitive tendency to imbue humanlike mental states to nonhuman entities (i.e., anthropomorphism). The present research tested whether anthropomorphizing slot machines would increase gambling. Four studies manipulated slot machine anthropomorphization and found that exposing people to an anthropomorphized description of a slot machine increased gambling behavior and reduced gambling outcomes. Such findings emerged using tasks that focused on gambling behavior (Studies 1 to 3) as well as in experimental paradigms that included gambling outcomes (Studies 2 to 4). We found that gambling outcomes decrease because participants primed with the anthropomorphic slot machine gambled more (Study 4). Furthermore, we found that high-arousal positive emotions (e.g., feeling excited) played a role in the effect of anthropomorphism on gambling behavior (Studies 3 and 4). Our research indicates that the psychological process of gambling-machine anthropomorphism can be advantageous for the gaming industry; however, this may come at great expense for gamblers' (and their families') economic resources and psychological well-being. (c) 2015 APA, all rights reserved).

  13. Slot Parameter Optimization for Multiband Antenna Performance Improvement Using Intelligent Systems

    Directory of Open Access Journals (Sweden)

    Erdem Demircioglu

    2015-01-01

    Full Text Available This paper discusses bandwidth enhancement for multiband microstrip patch antennas (MMPAs using symmetrical rectangular/square slots etched on the patch and the substrate properties. The slot parameters on MMPA are modeled using soft computing technique of artificial neural networks (ANN. To achieve the best ANN performance, Particle Swarm Optimization (PSO and Differential Evolution (DE are applied with ANN’s conventional training algorithm in optimization of the modeling performance. In this study, the slot parameters are assumed as slot distance to the radiating patch edge, slot width, and length. Bandwidth enhancement is applied to a formerly designed MMPA fed by a microstrip transmission line attached to the center pin of 50 ohm SMA connecter. The simulated antennas are fabricated and measured. Measurement results are utilized for training the artificial intelligence models. The ANN provides 98% model accuracy for rectangular slots and 97% for square slots; however, ANFIS offer 90% accuracy with lack of resonance frequency tracking.

  14. Chest radiography after minor chest trauma

    Energy Technology Data Exchange (ETDEWEB)

    Rossen, B.; Laursen, N.O.; Just, S.

    The results of chest radiography in 581 patients with blunt minor thoracic trauma were reviewed. Frontal and lateral views of the chest indicated pathology in 72 patients (12.4%). Pneumothorax was present in 16 patients; 4 had hemothorax. The physical examination and the results of chest radiography were not in accordance because in 6(30%) of the 20 patients with hemo/-pneumothorax the physical examination was normal. Consequently there is wide indication for chest radiography after minor blunt chest trauma.

  15. Chest radiography after minor chest trauma

    International Nuclear Information System (INIS)

    Rossen, B.; Laursen, N.O.; Just, S.

    1987-01-01

    The results of chest radiography in 581 patients with blunt minor thoracic trauma were reviewed. Frontal and lateral views of the chest indicated pathology in 72 patients (12.4%). Pneumothorax was present in 16 patients; 4 had hemothorax. The physical examination and the results of chest radiography were not in accordance because in 6(30%) of the 20 patients with hemo/-pneumothorax the physical examination was normal. Consequently there is wide indication for chest radiography after minor blunt chest trauma. (orig.)

  16. Early detection of drug-induced pneumonitis by gallium-67 lung scan in six patients with normal chest radiographs

    Energy Technology Data Exchange (ETDEWEB)

    Nakajima, H; Sawa, H; Takashima, S [Osaka City Univ. (Japan). Hospital

    1981-06-01

    Increased pulmonary accumulation of Gallium-67-citrate was observed in 6 patients (4 with malignant lymphoma, 1 with uterine cancer and 1 with acute myelocytic leukemia) preceding the appearance of any abnormal findings in both chest X-ray and blood gas data. All of them had received multiple courses of chemotherapy. In these patients, the anticancer drugs were administered for 13 to 22 weeks (mean 15 weeks). One patient with malignant lymphoma showed abnormal /sup 67/Ga lung uptake greater than hepatic activity, 3 patients (malignant lymphoma, 2 and uterine cancer, 1) visualized abnormal /sup 67/Ga lung uptake equal to hepatic activity and 2 cases (malignant lymphoma, 1 and acute myelocytic leukemia, 1) demonstrated abnormal accumulation of /sup 67/Ga in the lung greater than background activity. In 4 patients (3 with malignant lymphoma and 1 with uterine cancer) out of 6, transbronchial lung biopsy obtained after the /sup 67/Ga scans showed non-specific interstitial pneumonitis with infiltration of lymphocytes and macrophages compatible with drug-induced pneumonitis. In the other 2 patients, cytology and cultures were negative and follow up /sup 67/Ga lung scans revealed a reduction in intensity of uptake after treatment with corticosteroid. Therefore, we considered that the /sup 67/Ga lung scan was useful for early detection of drug-induced pneumonitis.

  17. Transmission dynamic range in chest radiology

    International Nuclear Information System (INIS)

    Lemmers, H.E.A.S.J.; Schultze Kool, L.J.; van Elburg, H.J.; Boelens, F.

    1989-01-01

    Due to the large difference in transmission between the lung area and the mediastinum, the human chest is a challenging object for radiographic imaging. This study is performed in order to define the dynamic range needed for a chest imaging chain. Eight hundred seventy-five consecutive outpatients were imaged with a prototype AMBER (advanced multiple beam equalization radiography) unit at 141 kVp. The equalization facility was disabled, allowing for the simultaneous capture of a film image and a digital dataset representing the local patient transmission in fields of approximately 2x2 cm. The datasets were analyzed to obtain the relation between the average transmission distribution in a subset of the population and physical parameters characterizing this subset, such as body weight or length

  18. Digital and conventional radiology techniques: comparison of dosage and costs

    International Nuclear Information System (INIS)

    Arranza, L.; Albornoz, C. de

    1996-01-01

    To compare the radiation dosage and costs in conventional and digital technologies. The study dealt with transverse sections. The dosage applied with conventional technology was measured in 254 patients who intertwined 402 explorations of 6 anatomic regions in 4 Radiodiagnostic Services. The dosage applied with digital technology was measured in 57 patients who underwent 95 explorations of the same anatomic region in one Radiodiagnostic Service. The costs of the 6 types of conventional and digital explorations performed were calculated for two Radiodiagnostic Service. The doses administered (mGy) using convectional/digital technology were as follows: chest PA 0.2/0.1; chest LAT 0.7/0.3; breast CC 7.0/8.4; breast LAT 7.0/7.8; breast OB 7.0/10.5; cervical spine AP 9.6/9.0; cervical spine LAT 21.9/29.6; pelvis AP 7.3/7.1; plain abdominal 6.5/2.2. The costs incurred (1992 pesetas) with the convectional/digital technologies: chest AP and LAT 1,393/2,973; portable chest 2,027/3,714; mammography 2,357/3,486; phlebography 12,718/14,023; hysterosalpingography 4,876/6,701; bone scientigraphy 1,633/2,839. Compared with conventional technology, digital imaging reduces the radiation doses received by the patients, except in the case of mammography. The costs associated with the use of digital technology are greater than those incurred with conventional technology, mainly due to the costs of amortization. the use of digital technology is more justified when: 1) it is very necessary to reduce the dosage; 2) studies of chest and abdomen predominant; 3) the volume of utilization is high; 4) staff management is flexible , and 5) the cost of purchasing the equipment is lower. (Author) 10 refs

  19. Effects of geometry on slot-jet film cooling performance

    Energy Technology Data Exchange (ETDEWEB)

    Hyams, D.G.; McGovern, K.T.; Leylek, J.H. [Clemson Univ., SC (United States)

    1995-10-01

    The physics of the film cooling process for shaped, inclined slot-jets with realistic slot-length-to-width ratios (L/s) is studied for a range of blowing ratio (M) and density ratio (DR) parameters typical of gas turbine operations. For the first time in the open literature, the effect of inlet and exit shaping of the slot-jet on both flow and thermal field characteristics is isolated, and the dominant mechanisms responsible for differences in these characteristics are documented. A previously documented computational methodology was applied for the study of four distinct configurations: (1) slot with straight edges and sharp corners (reference case); (2) slot with shaped inlet region; (3) slot with shaped exit region; and (4) slot with both shaped inlet and exit regions. Detailed field results as well as surface phenomena involving adiabatic film effectiveness ({eta}) and heat transfer coefficient (h) are presented. It is demonstrated that both {eta} and h results are vital in the proper assessment of film cooling performance. All simulations were carried out using a multi-block, unstructured/adaptive grid, fully explicit, time-marching solver with multi-grid, local time stepping, and residual smoothing type acceleration techniques. Special attention was paid to and full documentation provided for: (1) proper modeling of the physical phenomena; (2) exact geometry and high quality grid generation techniques; (3) discretization schemes; and (4) turbulence modeling issues. The key parameters M and DR were varied from 1.0 to 2.0 and 1.5 to 2.0, respectively, to show their influence. Simulations were repeated for slot length-to-width ratio (L/s) of 3.0 and 4.5 in order to explain the effects of this important parameter. Additionally, the performance of two popular turbulence models, standard k-F, and RNG k-E, were studied to establish their ability to handle highly elliptic jet/crossflow interaction type processes.

  20. Flexural and Thermal Properties of Novel Energy Conservation Slotted Reinforced Concrete Beams

    Directory of Open Access Journals (Sweden)

    Gao Ma

    2016-01-01

    Full Text Available Conventional solid reinforced concrete (RC beams were modified to slotted beams for consideration as thermal insulation structural components. The slotted beam consisted of an outer and an inner beam, respectively, with a slot located near the middle of the beam along its width direction for filling thermal insulation material. Flexural and thermal behavior of the slotted beams were investigated. Three RC reference solid beams and six slotted beams were fabricated and tested under four-point bending tests. The test results indicated that the failure mode of both slotted beams and the solid beams was flexural failure. However, the damage process of the slotted beams was different from that of the solid beams at the final loading stage. The moment curvature analysis indicated that the tensile reinforcement ratio of the outer and inner beams had an important effect on the flexural behavior, especially the ductility of the slotted beams. Thermal study indicated that the heat transfer coefficient of the slotted beam was greatly reduced and the thermal inertia factor increased a lot, compared with the solid beam. In addition, FE simulation results showed that a new frame structure using slotted beams exhibited obvious and attractive thermal insulation property.

  1. Routine chest and abdominal high-pitch CT: An alternative low dose protocol with preserved image quality

    International Nuclear Information System (INIS)

    Amacker, Nadja A.; Mader, Caecilia; Alkadhi, Hatem; Leschka, Sebastian; Frauenfelder, Thomas

    2012-01-01

    Objective: To investigate the radiation dose and image quality of the high-pitch dual source computer tomography (DSCT) for routine chest and abdominal scans. Methods: 130 consecutive patients (62 female, 68 male, median age 55 years) were included. All patients underwent 128-slice high-pitch DSCT (chest n = 99; abdomen n = 84) at a pitch of 3.2. Two observers independently rated image quality using a 4-point score (1: excellent to 4: non-diagnostic). Image noise was measured and operational radiation dose quantities were recorded. An additional group of 132 patients (chest, n = 80; abdomen n = 52) scanned with standard-pitch CT matched for age, gender, and body mass index (BMI) served as control group. Results: Interobserver agreement for image quality rating was good (k = 0.74). Subjective image quality of high-pitch CT was diagnostic in all patients (median score chest; 2, median score abdomen: 2). Image noise of high-pitch CT was comparable to standard-pitch for the chest (p = 0.32) but increased in the abdomen (p < 0.0001). For high-pitch CT radiation dose was 4.4 ± 0.9 mSv (chest) and 6.5 ± 1.2 mSv (abdomen). These values were significantly lower compared to standard-pitch CT (chest: 5.5 ± 1.2 mSv; abdomen: 11.3 ± 3.8 mSv). Conclusion: Based on the technical background high-pitch dual source CT may serve as an alternative scan mode for low radiation dose routine chest and abdominal CT.

  2. Sandstorm in the chest?

    Directory of Open Access Journals (Sweden)

    Talluri MR

    2011-07-01

    Full Text Available A 32 year old female presented with dry cough and progressive breathlessness of one year duration. There was no history suggestive of collagen vascular disease, lung parenchymal infection or allergic airway disease. Clinical evaluation showed basal fine inspiratory crepitations. Radiographic examination of the chest revealed a black pleura line and lung parenchymal calcification. CT scan of the chest demonstrated nodular calcification of lung parenchyma with a “crazy pavement” pattern, which is suggestive of alveolar calcification. Pulmonary function test showed a severe restrictive defect. On transbronchial lung biopsy calcific spherules suggestive of the alveolar microlithiasis were seen. Diagnosis of pulmonary alveolar microlithiasis was made and symptomatic treatment was given, as there is no specific therapy available. The case illustrates an unusual cause of shortness of breath in a young female with striking radiographic features.

  3. Influence of MR imaging in radiation therapy of chest lymphoma

    International Nuclear Information System (INIS)

    Carlsen, S.E.; Hoppe, R.; Bergin, C.J.

    1991-01-01

    This paper evaluates the influence of MR detection of additional sites of chest lymphoma on radiation therapy. Chest MR images and CT scans of 56 patients with new or recurrent mediastinal lymphoma obtained within 1 month of each other were retrospectively reviewed. MR images included T1- and T2-weighted SE and STIR sequences. Images were assessed for pleural and extrapleural disease. Radiation portals of patients with pleural or chest wall disease were reevaluated and compared with portals originally designed with CT. MR imaging demonstrated chest wall disease in 15 patients (21 sites). Ten patients also had pleural disease (13 sites). CT identified chest wall disease in four of these patients (five sites) and pleural disease in three patients (five sites). Seven of the 15 patients with chest wall disease were treated with radiation therapy alone. Two of the seven patients had significant modification of radiation portals based on MR findings. Retrospectively, therapy would have been altered in an additional two patients in whom pleural disease was identified at MR. The increased sensitivity of MR in detecting chest wall or pleural disease has important implications for treatment planning in chest wall lymphoma

  4. Evaluation of digital dental models obtained from dental cone-beam computed tomography scan of alginate impressions

    Science.gov (United States)

    Jiang, Tingting; Lee, Sang-Mi; Hou, Yanan; Chang, Xin

    2016-01-01

    Objective To investigate the dimensional accuracy of digital dental models obtained from the dental cone-beam computed tomography (CBCT) scan of alginate impressions according to the time elapse when the impressions are stored under ambient conditions. Methods Alginate impressions were obtained from 20 adults using 3 different alginate materials, 2 traditional alginate materials (Alginoplast and Cavex Impressional) and 1 extended-pour alginate material (Cavex ColorChange). The impressions were stored under ambient conditions, and scanned by CBCT immediately after the impressions were taken, and then at 1 hour intervals for 6 hours. After reconstructing three-dimensional digital dental models, the models were measured and the data were analyzed to determine dimensional changes according to the elapsed time. The changes within the measurement error were regarded as clinically acceptable in this study. Results All measurements showed a decreasing tendency with an increase in the elapsed time after the impressions. Although the extended-pour alginate exhibited a less decreasing tendency than the other 2 materials, there were no statistically significant differences between the materials. Changes above the measurement error occurred between the time points of 3 and 4 hours after the impressions. Conclusions The results of this study indicate that digital dental models can be obtained simply from a CBCT scan of alginate impressions without sending them to a remote laboratory. However, when the impressions are not stored under special conditions, they should be scanned immediately, or at least within 2 to 3 hours after the impressions are taken. PMID:27226958

  5. Analysis and design of broadband U-slot cut rectangular microstrip ...

    Indian Academy of Sciences (India)

    AMIT A DESHMUKH

    2017-07-15

    Jul 15, 2017 ... Abstract. Broadband microstrip antenna using variations of U-slot has been widely reported. However, in most of the reported work, an in-depth explanation about the mode introduced by U-slot and procedure to design. U-slot cut antennas at any given frequency is not explained. In this paper, first an ...

  6. Chest Tomosynthesis: Technical Principles and Clinical Update

    Science.gov (United States)

    Dobbins, James T.; McAdams, H. Page

    2009-01-01

    Digital tomosynthesis is a radiographic technique that can produce an arbitrary number of section images of a patient from a single pass of the x-ray tube. It utilizes a conventional x-ray tube, a flat-panel detector, a computer-controlled tube mover, and special reconstruction algorithms to produce section images. While it does not have the depth resolution of computed tomography (CT), tomosynthesis provides some of the tomographic benefits of CT but at lower cost and radiation dose than CT. Compared to conventional chest radiography, chest tomosynthesis results in improved visibility of normal structures such as vessels, airway and spine. By reducing visual clutter from overlying normal anatomy, it also enhances detection of small lung nodules. This review article outlines the components of a tomosynthesis system, discusses results regarding improved lung nodule detection from the recent literature, and presents examples of nodule detection from a clinical trial in human subjects. Possible implementation strategies for use in clinical chest imaging are discussed. PMID:19616909

  7. Two cases of Chest Heating Sensation treated by Hwangryunhaedok-tang Herbal-Acupuncture

    Directory of Open Access Journals (Sweden)

    Gwon-Il Cho

    2003-06-01

    Full Text Available The purpose of this study is to evaluate the clinical effect of Hwangryunhaedok-tang Herbal-Acupuncture. Hwangryunhaedok-tang is used in all heating diseases. Chest Heating Sensation is a unique concept in Oriental Medicine. So we applied Hwangryunhaedok-tang Herbal-Acupuncture to treat the Chest Heating Sensation. We used DITI(Digital Infrared Thermographic Imaging to estimate the temperatures of chest surface for the outcome assessment. We came to know that the chest surface temperatures were all reduced in both cases after Herbal-Acupuncture treatment. The reduced average temperature was 1.5℃ in case 1 and 0.9℃ in case 2. The above result indicates that Hwangryunhaedok-tang Herbal-Acupuncture treatment has an effect on Chest Heating Sensation, thus continuous Hwangryunhaedok-tang Herbal-Acupuncture study will be needed for more clinical applications.

  8. Simultaneous measurement of refractive index and thickness distributions using low-coherence digital holography and vertical scanning

    International Nuclear Information System (INIS)

    Watanabe, Kaho; Ohshima, Masashi; Nomura, Takanori

    2014-01-01

    The simultaneous measurement method of a refractive index distribution and a thickness distribution using low-coherence digital holography with a vertical scanning is proposed. The proposed method consists of a combination of digital holography and low-coherence interferometry. The introduction of a datum plane enables the measurement of both a refractive index distribution and a thickness distribution. By the optical experiment, the potential of the proposed method is confirmed. (paper)

  9. Digital vs conventional radiography: cost and revenue analysis

    International Nuclear Information System (INIS)

    Dalla Palma, L.; Cuttin, R.; Rimondini, A.; Grisi, G.

    1999-01-01

    The objective of this study was to analyse and compare the operating and investment costs of two radiographic systems, a conventional and a digital one, and to evaluate the cost/revenue ratio of the two systems. The radiological activity over 1 year for chest and skeletal exams was evaluated: 13,401 chest and 7,124 skeletal exams were considered. The following parameters of variable costs were evaluated: the difference between variable proportional costs of the two technologies, the effective variable cost of any size film, including the chemicals, and for different sizes of digital film, variable costs of chest plus skeletal exams performed with the two techniques. Afterwards the economical effect was considered taking into account depreciation during a time of utilization ranging between 8 and 4 years. In the second part of the analysis the total cost and the revenues of the two technologies were determined. The comparison between the digital and conventional systems has shown the following aspects: 1. Digital radiography system has a much higher investment cost in comparison with the conventional one. 2. Operating costs of digital equipment are higher or lower depending on the film size used. Evaluating chest X-ray we reach a breakeven point after 1 year and 10,000 exams only if displayed over 8 x 10-in. film and after 30,000 if displayed over a 11 x 14-in. film. 3. The total cost (variable cost, technology cost, labour cost) of digital technology is lower than that of the conventional system by 20 % on average using 8 x 10-in. film size. 4. Digital technology also allows lesser film waste and lesser film per exam (orig.)

  10. Estimated risk of radiation-induced cancer from paediatric chest CT: two-year cohort study

    Energy Technology Data Exchange (ETDEWEB)

    Niemann, Tilo [Cantonal Hospital Baden, Department of Radiology, Baden (Switzerland); University Lille Nord de France, Department of Thoracic Imaging, Hospital Calmette, Lille (France); Colas, Lucie; Santangelo, Teresa; Faivre, Jean Baptiste; Remy, Jacques; Remy-Jardin, Martine [University Lille Nord de France, Department of Thoracic Imaging, Hospital Calmette, Lille (France); Roser, Hans W.; Bremerich, Jens [University of Basel Hospital, Clinic of Radiology and Nuclear Medicine, Medical Physics, Basel (Switzerland)

    2015-03-01

    The increasing absolute number of paediatric CT scans raises concern about the safety and efficacy and the effects of consecutive diagnostic ionising radiation. To demonstrate a method to evaluate the lifetime attributable risk of cancer incidence/mortality due to a single low-dose helical chest CT in a two-year patient cohort. A two-year cohort of 522 paediatric helical chest CT scans acquired using a dedicated low-dose protocol were analysed retrospectively. Patient-specific estimations of radiation doses were modelled using three different mathematical phantoms. Per-organ attributable cancer risk was then estimated using epidemiological models. Additional comparison was provided for naturally occurring risks. Total lifetime attributable risk of cancer incidence remains low for all age and sex categories, being highest in female neonates (0.34%). Summation of all cancer sites analysed raised the relative lifetime attributable risk of organ cancer incidence up to 3.6% in female neonates and 2.1% in male neonates. Using dedicated scan protocols, total lifetime attributable risk of cancer incidence and mortality for chest CT is estimated low for paediatric chest CT, being highest for female neonates. (orig.)

  11. Black Lung Benefits Act: standards for chest radiographs. Final rule.

    Science.gov (United States)

    2014-04-17

    Physicians and adjudicators use chest radiographs (X-rays) as a tool in evaluating whether a coal miner suffers from pneumoconiosis (black lung disease). Accordingly, the Department's regulations implementing the Black Lung Benefits Act allow the submission of radiographs in connection with benefit claims and set out quality standards for administering and interpreting film-based chest radiographs. This final rule updates the Department's existing film-radiograph standards and provides parallel standards for digital radiographs. This rule also updates outdated terminology and removes certain obsolete provisions.

  12. Posterior diaphragmatic defect detected on chest CT: the incidence according to age and the lateral chest radiographic appearances

    International Nuclear Information System (INIS)

    Lee, Son Youl; Choi, Yo Won; Jeon, Seok Chol; Heo, Jeong Nam; Park, Choong Ki

    2007-01-01

    We wanted to investigate the incidence of posterior diaphragmatic defect on chest CT in various age groups and its lateral chest radiographic appearances. The chest CT scans of 78 patients of various ages with posterior diaphragmatic defect were selected among 1,991 patients, and they were analyzed for the incidence of defect in various age groups, the defect location and the herniated contents. Their lateral chest radiographs were analyzed for the shape of the posterior diaphragm and the posterior costophrenic sulcus. The patients' ages ranged from 34 to 87 with the tendency of a higher incidence in the older patients. The defect most frequently involved the medial two thirds (n = 49, 50.4%) and middle one third (n = 36, 37%) of the posterior diaphragm. The retroperitoneal fat was herniated into the thorax through the defect in all patients, and sometimes with the kidney (n = 8). Lateral chest radiography showed a normal diaphragmatic contour (n = 51, 49.5%), blunting of the posterior costophrenic sulcus (n = 41, 39.8%), focal humping of the posterior diaphragm (n = 7, 6.8%), or upward convexity (n = 4, 3.9%) of the posterior costophrenic sulcus on the affected side. The posterior diaphragmatic defect discovered in asymptomatic patients who are without a history of peridiaphragmatic disease is most likely acquired, and this malady increases in incidence according to age. An abnormal contour of the posterior diaphragm or the costophrenic sulcus on a lateral chest radiograph may be a finding of posterior diaphragmatic defect

  13. Novel Influenza A (H1N1) Virus Infection in Children: Chest Radiographic and CT Evaluation

    Energy Technology Data Exchange (ETDEWEB)

    Choi, Min Jeong; Lee, Young Seok; Lee, Jee Young; Lee, Kun Song [Dankook University College of Medicine, Dankook University Hospital, Cheonan (Korea, Republic of)

    2010-12-15

    The purpose of this study was to evaluate the chest radiographic and CT findings of novel influenza A (H1N1) virus infection in children, the population that is more vulnerable to respiratory infection than adults. The study population comprised 410 children who were diagnosed with an H1N1 infection from August 24, 2009 to November 11, 2009 and underwent chest radiography at Dankook University Hospital in Korea. Six of these patients also underwent chest CT. The initial chest radiographs were classified as normal or abnormal. The abnormal chest radiographs and high resolution CT scans were assessed for the pattern and distribution of parenchymal lesions, and the presence of complications such as atelectasis, pleural effusion, and pneumomediastinum. The initial chest radiograph was normal in 384 of 410 (94%) patients and abnormal in 26 of 410 (6%) patients. Parenchymal abnormalities seen on the initial chest radiographs included prominent peribronchial marking (25 of 26, 96%), consolidation (22 of 26, 85%), and ground-glass opacities without consolidation (2 of 26, 8%). The involvement was usually bilateral (19 of 26, 73%) with the lower lung zone predominance (22 of 26, 85%). Atelectasis was observed in 12 (46%) and pleural effusion in 11 (42%) patients. CT (n = 6) scans showed peribronchovascular interstitial thickening (n = 6), ground-glass opacities (n = 5), centrilobular nodules (n = 4), consolidation (n = 3), mediastinal lymph node enlargement (n = 5), pleural effusion (n = 3), and pneumomediastinum (n = 3). Abnormal chest radiographs were uncommon in children with a swine-origin influenza A (H1N1) virus (S-OIV) infection. In children, H1N1 virus infection can be included in the differential diagnosis, when chest radiographs and CT scans show prominent peribronchial markings and ill-defined patchy consolidation with mediastinal lymph node enlargement, pleural effusion and pneumomediastinum

  14. Novel Influenza A (H1N1) Virus Infection in Children: Chest Radiographic and CT Evaluation

    International Nuclear Information System (INIS)

    Choi, Min Jeong; Lee, Young Seok; Lee, Jee Young; Lee, Kun Song

    2010-01-01

    The purpose of this study was to evaluate the chest radiographic and CT findings of novel influenza A (H1N1) virus infection in children, the population that is more vulnerable to respiratory infection than adults. The study population comprised 410 children who were diagnosed with an H1N1 infection from August 24, 2009 to November 11, 2009 and underwent chest radiography at Dankook University Hospital in Korea. Six of these patients also underwent chest CT. The initial chest radiographs were classified as normal or abnormal. The abnormal chest radiographs and high resolution CT scans were assessed for the pattern and distribution of parenchymal lesions, and the presence of complications such as atelectasis, pleural effusion, and pneumomediastinum. The initial chest radiograph was normal in 384 of 410 (94%) patients and abnormal in 26 of 410 (6%) patients. Parenchymal abnormalities seen on the initial chest radiographs included prominent peribronchial marking (25 of 26, 96%), consolidation (22 of 26, 85%), and ground-glass opacities without consolidation (2 of 26, 8%). The involvement was usually bilateral (19 of 26, 73%) with the lower lung zone predominance (22 of 26, 85%). Atelectasis was observed in 12 (46%) and pleural effusion in 11 (42%) patients. CT (n = 6) scans showed peribronchovascular interstitial thickening (n = 6), ground-glass opacities (n = 5), centrilobular nodules (n = 4), consolidation (n = 3), mediastinal lymph node enlargement (n = 5), pleural effusion (n = 3), and pneumomediastinum (n = 3). Abnormal chest radiographs were uncommon in children with a swine-origin influenza A (H1N1) virus (S-OIV) infection. In children, H1N1 virus infection can be included in the differential diagnosis, when chest radiographs and CT scans show prominent peribronchial markings and ill-defined patchy consolidation with mediastinal lymph node enlargement, pleural effusion and pneumomediastinum

  15. Examination of the constancy of the medium optical density of conventional chest radiographs

    International Nuclear Information System (INIS)

    Stieve, F.E.; Rupp, N.; Niederhofer, F.; Gfirtner, H.

    2002-01-01

    Purpose: To determine whether the medium optical density of conventional chest radiographs is constant. Materials and Methods: 47 PA chest radiographs were digitized to measure the optical density and its dynamic range of normal and pathologically altered lungs. Results: For constant exposure factors and a correctly positioned region of interest, the medium optical density deviates only minimal despite differences in AP chest diameter, entry dose and dynamic range, and is largely independent of the pathologic pulmonary findings. Conclusion: Digitization of the radiograph enables the determination of the physical parameters used to operate the imaging system, such as medium optical density, image contrast, image latitude and density distribution, in accordance to the guidelines of the German Medical Association. If the region of interest encompasses the clinically relevant region of the lung, the measured value of the optical density corresponds with only minor variation to the average value of the entire lung region. (orig.) [de

  16. Demonstration of slot-waveguide structures on silicon nitride / silicon oxide platform.

    Science.gov (United States)

    Barrios, C A; Sánchez, B; Gylfason, K B; Griol, A; Sohlström, H; Holgado, M; Casquel, R

    2007-05-28

    We report on the first demonstration of guiding light in vertical slot-waveguides on silicon nitride/silicon oxide material system. Integrated ring resonators and Fabry-Perot cavities have been fabricated and characterized in order to determine optical features of the slot-waveguides. Group index behavior evidences guiding and confinement in the low-index slot region at O-band (1260-1370nm) telecommunication wavelengths. Propagation losses of <20 dB/cm have been measured for the transverse-electric mode of the slot-waveguides.

  17. Detection of tuberculosis patterns in digital photographs of chest X-ray images using Deep Learning: feasibility study.

    Science.gov (United States)

    Becker, A S; Blüthgen, C; Phi van, V D; Sekaggya-Wiltshire, C; Castelnuovo, B; Kambugu, A; Fehr, J; Frauenfelder, T

    2018-03-01

    To evaluate the feasibility of Deep Learning-based detection and classification of pathological patterns in a set of digital photographs of chest X-ray (CXR) images of tuberculosis (TB) patients. In this prospective, observational study, patients with previously diagnosed TB were enrolled. Photographs of their CXRs were taken using a consumer-grade digital still camera. The images were stratified by pathological patterns into classes: cavity, consolidation, effusion, interstitial changes, miliary pattern or normal examination. Image analysis was performed with commercially available Deep Learning software in two steps. Pathological areas were first localised; detected areas were then classified. Detection was assessed using receiver operating characteristics (ROC) analysis, and classification using a confusion matrix. The study cohort was 138 patients with human immunodeficiency virus (HIV) and TB co-infection (median age 34 years, IQR 28-40); 54 patients were female. Localisation of pathological areas was excellent (area under the ROC curve 0.82). The software could perfectly distinguish pleural effusions from intraparenchymal changes. The most frequent misclassifications were consolidations as cavitations, and miliary patterns as interstitial patterns (and vice versa). Deep Learning analysis of CXR photographs is a promising tool. Further efforts are needed to build larger, high-quality data sets to achieve better diagnostic performance.

  18. Reading a radiologist's mind: monitoring rising and falling interest levels while scanning chest x-rays

    Science.gov (United States)

    Alzubaidi, Mohammad; Patel, Ameet; Panchanathan, Sethuraman; Black, John A., Jr.

    2010-02-01

    Radiological images constitute a special class of images that are captured (or computed) specifically for the purpose of diagnosing patients. However, because these are not "natural" images, radiologists must be trained to interpret them through a process called "perceptual learning". However, because perceptual learning is implicit, experienced radiologists may sometimes find it difficult to explicitly (i.e. verbally) train less experienced colleagues. As a result, current methods of training can take years before a new radiologist is fully competent to independently interpret medical images. We hypothesize that eye tracking technology (coupled with multimedia technology) can be used to accelerate the process of perceptual training, through a Hebbian learning process. This would be accomplished by providing a radiologist-in-training with real-time feedback as he/she is fixating on important regions of an image. Of course this requires that the training system have information about what regions of an image are important - information that could presumably be solicited from experienced radiologists. However, our previous work has suggested that experienced radiologists are not always aware of those regions of an image that attract their attention, but are not clinically significant - information that is very important to a radiologist in training. This paper discusses a study in which local entropy computations were done on scan path data, and were found to provide a quantitative measure of the moment-by-moment interest level of radiologists as they scanned chest x-rays. The results also showed a striking contrast between the moment-by-moment deployment of attention between experienced radiologists and radiologists in training.

  19. FDTD Analysis of U-Slot Rectangular Patch Antenna

    Science.gov (United States)

    Luk, K. M.; Tong, K. F.; Shum, S. M.; Lee, K. F.; Lee, R. Q.

    1997-01-01

    The U-slot rectangular patch antenna (Figure I) has been found experimentally to provide impedance and gain bandwidths of about 300 without the need of stacked or coplanar parasitic elements [1,2]. In this paper, simulation results of the U-slot patch using FDTD analysis are presented. Comparison with measured results are given.

  20. Possibilities of CT examinations by chest trauma

    International Nuclear Information System (INIS)

    Ftacnikova, B.

    1994-01-01

    Chest trauma represents the most frequent associated injury in multiply injured patients. The success of treatment depends also on prompt and effective diagnosis and extent of the injuries, on quality interdisciplinary approach. Author presents contributions of computed tomography (CT) in the management of 77 critically injured patients. Attention is focused on the efficacy of CT examination routinely employed in the setting of thoracic trauma and its relationship to following rationalization of treatment. CT scans of thorax is modality of choice for evaluating patients with occult pneumothorax, chest wall deformity of rib fractures, early diagnosis of lung contusion and laceration. (author). 13 figs., 1 tab., 7 refs

  1. The chest radiological manifestation in psittacosis

    International Nuclear Information System (INIS)

    Zhang Jun; Fu Jiazhen; Wang Shulan; Zhang Shuxin; Sun Guochang; Tang Guangjian

    2005-01-01

    Objective: To summarize the clinical characteristics and imaging features of psittacosis. Methods: The clinical features and imaging appearances of 3 cases with acute psittacosis were retrospectively analyzed. The related literature was reviewed. Results: The clinical manifestation of psittacosis was high fever in the patients. Physical findings included pulse-temperature dissociation, localized lung crackles, hepatomegaly, and splenomegaly. Laboratory findings showed elevation of ESR in all cases, and liver dysfunction was present in 2 cases. The counts of white blood cells were normal, but the percent of neutrophils might be increased. The chest X-ray and CT scan showed air-space consolidation and ground-glass attenuation in the lung, and miliary, nodular, or consolidated shadows were found in pathological areas. Pleural effusions were also present in 2 cases. Psittacosis was diagnosed from the history of exposure to infected parrots and elevation of the IgG and IgM titer for Chlamydia psittaci. Erythromycin was effective in all 3 patients. Conclusion: Although the appearance of psittacosis on clinical findings and chest X-ray and CT scan is not characteristic, psittacosis can be diagnosed with the combination of the history of exposure to infected parrots and laboratory findings. CT scan can reveal the focus earlier and accurately, and catching the imaging features of psittacosis is helpful in differential diagnosis. (authors)

  2. Doses in pediatric patients undergoing chest and abdomen CT examinations. Preliminary results

    International Nuclear Information System (INIS)

    Jornada, Tiago S.; Silva, Teogenes A. da

    2011-01-01

    Computed tomography (CT) is a non-invasive method of image production that imparts significant doses to a patient, it is expected that pediatric CT examinations will increase the risk of induced cancer in children. In this study the effective doses in a five year-old child submitted to chest or abdomen CT scans were assessed for comparison purposes. The CTEXPO computed program was used with data from routine protocols of a 0 to 13 year-old children in two public hospitals in Belo Horizonte. Hospital A used a Siemens Dual-Slice unit with 80 kV, 41 mA and pitch 2 for chest or abdomen; hospital B used a Multislice GE unit with 120 kV, 45 mA and pitch 1 for chest and 120 kV, 55 mA. and pitch 1 for abdomen. Results of effective doses in a five year-old child were 1.7 and 1.0 mSv in hospital A and 9.1 and 7.2 mSv in hospital B, for chest and abdomen, respectively. Results were compared to the reference effective doses of 7.2 and 5.0 mSv for chest and abdomen respectively that were derived from the air kerma length product values given in ICRP publication 87. Results of hospital A showed that low dose exposures also can be achieved in CT scans of children. Results showed that even a hospital with a modern facility (hospital B) can provided doses higher than reference values if protocols are not adjusted for children. Preliminary results suggested that there is a room for optimizing children exposure submitted to CT scans. (author)

  3. TU-FG-209-11: Validation of a Channelized Hotelling Observer to Optimize Chest Radiography Image Processing for Nodule Detection: A Human Observer Study

    International Nuclear Information System (INIS)

    Sanchez, A; Little, K; Chung, J; Lu, ZF; MacMahon, H; Reiser, I

    2016-01-01

    Purpose: To validate the use of a Channelized Hotelling Observer (CHO) model for guiding image processing parameter selection and enable improved nodule detection in digital chest radiography. Methods: In a previous study, an anthropomorphic chest phantom was imaged with and without PMMA simulated nodules using a GE Discovery XR656 digital radiography system. The impact of image processing parameters was then explored using a CHO with 10 Laguerre-Gauss channels. In this work, we validate the CHO’s trend in nodule detectability as a function of two processing parameters by conducting a signal-known-exactly, multi-reader-multi-case (MRMC) ROC observer study. Five naive readers scored confidence of nodule visualization in 384 images with 50% nodule prevalence. The image backgrounds were regions-of-interest extracted from 6 normal patient scans, and the digitally inserted simulated nodules were obtained from phantom data in previous work. Each patient image was processed with both a near-optimal and a worst-case parameter combination, as determined by the CHO for nodule detection. The same 192 ROIs were used for each image processing method, with 32 randomly selected lung ROIs per patient image. Finally, the MRMC data was analyzed using the freely available iMRMC software of Gallas et al. Results: The image processing parameters which were optimized for the CHO led to a statistically significant improvement (p=0.049) in human observer AUC from 0.78 to 0.86, relative to the image processing implementation which produced the lowest CHO performance. Conclusion: Differences in user-selectable image processing methods on a commercially available digital radiography system were shown to have a marked impact on performance of human observers in the task of lung nodule detection. Further, the effect of processing on humans was similar to the effect on CHO performance. Future work will expand this study to include a wider range of detection/classification tasks and more

  4. Mechanism Design for Multi-slot Ads Auction in Sponsored Search Markets

    Science.gov (United States)

    Deng, Xiaotie; Sun, Yang; Yin, Ming; Zhou, Yunhong

    In this paper, we study pricing models for multi-slot advertisements, where advertisers can bid to place links to their sales webpages at one or multiple slots on a webpage, called the multi-slot AD auction problem. We develop and analyze several important mechanisms, including the VCG mechanism for multi-slot ads auction, the optimal social welfare solution, as well as two weighted GSP-like protocols (mixed and hybrid). Furthermore, we consider that forward-looking Nash equilibrium and prove its existence in the weighted GSP-like pricing protocols.

  5. The transverse diameter of the chest on routine radiographs reliably estimates gestational age and weight in premature infants.

    Science.gov (United States)

    Dietz, Kelly R; Zhang, Lei; Seidel, Frank G

    2015-08-01

    Prior to digital radiography it was possible for a radiologist to easily estimate the size of a patient on an analog film. Because variable magnification may be applied at the time of processing an image, it is now more difficult to visually estimate an infant's size on the monitor. Since gestational age and weight significantly impact the differential diagnosis of neonatal diseases and determine the expected size of kidneys or appearance of the brain by MRI or US, this information is useful to a pediatric radiologist. Although this information may be present in the electronic medical record, it is frequently not readily available to the pediatric radiologist at the time of image interpretation. To determine if there was a correlation between gestational age and weight of a premature infant with their transverse chest diameter (rib to rib) on admission chest radiographs. This retrospective study was approved by the institutional review board, which waived informed consent. The maximum transverse chest diameter outer rib to outer rib was measured on admission portable chest radiographs of 464 patients admitted to the neonatal intensive care unit (NICU) during the 2010 calendar year. Regression analysis was used to investigate the association between chest diameter and gestational age/birth weight. Quadratic term of chest diameter was used in the regression model. Chest diameter was statistically significantly associated with both gestational age (P chest diameter on digital chest radiograph with the tables and graphs in our study.

  6. The value of digital tomosynthesis of the chest as a problem-solving tool for suspected pulmonary nodules and hilar lesions detected on chest radiography

    Energy Technology Data Exchange (ETDEWEB)

    Galea, Angela, E-mail: galeaangie@gmail.com [Peninsula Radiology Academy, William Prance Road, Plymouth PL65WR (United Kingdom); Dubbins, Paul, E-mail: Paul.dubbins@nhs.net [Plymouth Hospital NHS Trust, Plymouth PL68DH (United Kingdom); Riordan, Richard, E-mail: richardriordan@nhs.net [Plymouth Hospital NHS Trust, Plymouth PL68DH (United Kingdom); Adlan, Tarig, E-mail: tarig.adlan@nhs.net [Plymouth Hospital NHS Trust, Plymouth PL68DH (United Kingdom); Roobottom, Carl, E-mail: carl.roobotoom@nhs.net [Plymouth Hospital NHS Trust, Plymouth PL68DH (United Kingdom); Gay, David, E-mail: davegay@nhs.net [Plymouth Hospital NHS Trust, Plymouth PL68DH (United Kingdom)

    2015-05-15

    Graphical abstract: When compared to CXR, DTS has: • Superior resolution • Better assessment of location in the AP dimension (better at locating a pleural or intrapulmonary lesion) • Better characterisation (better at distinguishing between calcified plaque and soft tissue) • Removes composite artefact caused by overlying anatomical structures (such as the ribs or pulmonary vessels) DTS has improved sensitivity, specificity and accuracy when compared to CXR. - Highlights: • DTS is a type of limited angle tomography. Sixty coronal reconstructed images of the chest are produced that combine the superior resolution of radiography with the tomographic benefits of computed tomography. • The sensitivity for detecting a suspected lung lesions is 0.65 with CXR and 0.91 for DTS. • The high specificity of DTS (1) and the high negative predictive value (0.94) are similar to CT and suggest that if the DTS is normal patients do not need further assessment with CT with significant potential dose savings. • 50% of suspected lesions were resolved with CXR, this improved to 96% with DTS. - Abstract: Objectives: To assess the capability of digital tomosynthesis (DTS) of the chest compared to a postero-anterior (PA) and lateral chest radiograph (CXR) in the diagnosis of suspected but unconfirmed pulmonary nodules and hilar lesions detected on a CXR. Computed tomography (CT) was used as the reference standard. Materials and method: 78 patients with suspected non-calcified pulmonary nodules or hilar lesions on their CXR were included in the study. Two radiologists, blinded to the history and CT, prospectively analysed the CXR (PA and lateral) and the DTS images using a picture archiving and communication workstation and were asked to designate one of two outcomes: true intrapulmonary lesion or false intrapulmonary lesion. A CT of the chest performed within 4 weeks of the CXR was used as the reference standard. Inter-observer agreement and time to report the modalities

  7. The value of digital tomosynthesis of the chest as a problem-solving tool for suspected pulmonary nodules and hilar lesions detected on chest radiography

    International Nuclear Information System (INIS)

    Galea, Angela; Dubbins, Paul; Riordan, Richard; Adlan, Tarig; Roobottom, Carl; Gay, David

    2015-01-01

    Graphical abstract: When compared to CXR, DTS has: • Superior resolution • Better assessment of location in the AP dimension (better at locating a pleural or intrapulmonary lesion) • Better characterisation (better at distinguishing between calcified plaque and soft tissue) • Removes composite artefact caused by overlying anatomical structures (such as the ribs or pulmonary vessels) DTS has improved sensitivity, specificity and accuracy when compared to CXR. - Highlights: • DTS is a type of limited angle tomography. Sixty coronal reconstructed images of the chest are produced that combine the superior resolution of radiography with the tomographic benefits of computed tomography. • The sensitivity for detecting a suspected lung lesions is 0.65 with CXR and 0.91 for DTS. • The high specificity of DTS (1) and the high negative predictive value (0.94) are similar to CT and suggest that if the DTS is normal patients do not need further assessment with CT with significant potential dose savings. • 50% of suspected lesions were resolved with CXR, this improved to 96% with DTS. - Abstract: Objectives: To assess the capability of digital tomosynthesis (DTS) of the chest compared to a postero-anterior (PA) and lateral chest radiograph (CXR) in the diagnosis of suspected but unconfirmed pulmonary nodules and hilar lesions detected on a CXR. Computed tomography (CT) was used as the reference standard. Materials and method: 78 patients with suspected non-calcified pulmonary nodules or hilar lesions on their CXR were included in the study. Two radiologists, blinded to the history and CT, prospectively analysed the CXR (PA and lateral) and the DTS images using a picture archiving and communication workstation and were asked to designate one of two outcomes: true intrapulmonary lesion or false intrapulmonary lesion. A CT of the chest performed within 4 weeks of the CXR was used as the reference standard. Inter-observer agreement and time to report the modalities

  8. Slotted Blades Savonius Wind Turbine Analysis by CFD

    Directory of Open Access Journals (Sweden)

    Andrea Alaimo

    2013-12-01

    Full Text Available In this paper a new bucket configuration for a Savonius wind generator is proposed. Numerical analyses are performed to estimate the performances of the proposed configuration by means of the commercial code COMSOL Multiphysics® with respect to Savonius wind turbine with overlap only. Parametric analyses are performed, for a fixed overlap ratio, by varying the slot position; the results show that for slot positioned near the blade root, the Savonius rotor improves performances at low tip speed ratio, evidencing a better starting torque. This circumstance is confirmed by static analyses performed on the slotted blades in order to investigate the starting characteristic of the proposed Savonius wind generator configuration.

  9. Slots in dielectric image line as mode launchers and circuit elements

    Science.gov (United States)

    Solbach, K.

    1981-01-01

    A planar resonator model is used to investigate slots in the ground plane of dielectric image lines. An equivalent circuit representation of the slot discontinuity is obtained, and the launching efficiency of the slot as a mode launcher is analyzed. Slots are also shown to be useful in the realization of dielectric image line array antennas. It is found that the slot discontinuity can be shown as a T-junction of the dielectric image line and a metal waveguide. The launching efficiency is found to increase with the dielectric constant of the dielectric image line, exhibiting a maximum value for guides whose height is slightly less than half a wavelength in the dielectric medium. The measured launching efficiencies of low permittivity dielectric image lines are found to be in good agreement with calculated values

  10. The Role of Digital 3D Scanned Models in Dental Students' Self-Assessments in Preclinical Operative Dentistry.

    Science.gov (United States)

    Lee, Cliff; Kobayashi, Hiro; Lee, Samuel R; Ohyama, Hiroe

    2018-04-01

    The aim of this study was to determine how dental student self-assessment and faculty assessment of operative preparations compared for conventional visual assessment versus assessment of scanned digital 3D models. In 2016, all third-year students in the Class of 2018 (N=35) at Harvard School of Dental Medicine performed preclinical exams of Class II amalgam preparations (C2AP) and Class III composite preparations (C3CP) and completed self-assessment forms; in 2017, all third-year students in the Class of 2019 (N=34) performed the same exams. Afterwards, the prepared typodont teeth were digitally scanned. Students self-assessed their preparations digitally, and four faculty members graded the preparations conventionally and digitally. The results showed that, overall, the students assessed their preparations higher than the faculty assessments. The mean student-faculty gaps for C2AP and C3CP in the conventional assessments were 11% and 5%, respectively. The mean digital student-faculty gap for C2AP and C3CP were 8% and 2%, respectively. In the conventional assessments, preclinical performance was negatively correlated with the student-faculty gap (r=-0.47, pStudents in the bottom quartile significantly improved their self-assessment accuracy using digital self-assessments over conventional assessments (C2AP 10% vs. 17% and C3CP 3% vs. 10%, respectively). These results suggest that digital assessments offered a significant learning opportunity for students to critically self-assess themselves in operative preclinical dentistry. The lower performing students benefitted the most, improving their assessment ability to the level of the rest of the class.

  11. Possibilities for exposure reduction in computed tomography examination of acute chest pain; Moeglichkeiten der Dosisreduktion bei CT-Untersuchungen des akuten Thoraxschmerzes

    Energy Technology Data Exchange (ETDEWEB)

    Becker, H.C. [Klinikum der Ludwig-Maximilians-Universitaet Muenchen, Campus Grosshadern, Institut fuer Klinische Radiologie, Muenchen (Germany)

    2012-10-15

    Electrocardiogram-gated (ECG) computed tomography (CT) investigations can be accompanied by high amounts of radiation exposure. This is particularly true for the investigation of patients with unclear and acute chest pain. The common approach in patients with acute chest pain is standard spiral CT of the chest. The chest pain or triple-rule-out CT protocol is a relatively new ECG-gated protocol of the entire chest. This article reviews and discusses different techniques for the CT investigation of patients with acute chest pain. By applying the appropriate scan technique, the radiation exposure for an ECG-gated protocol must not necessarily be higher than a standard chest CT scan Aortic pathologies are far better depicted by ECG-gated scan protocols and depending on the heart rate coronary artery disease can also be detected at the same time. The use of ECG-triggered scans will not support the diagnostics of the pulmonary arteries. However, in unspecific chest pain an ECG-triggered scan protocol can provide information on the differential diagnosis. (orig.) [German] EKG-getriggerte CT-Untersuchungen koennen mit einer relativ hohen Strahlenexposition einhergehen. Dies gilt im besonderen Masse fuer die Untersuchung des gesamten Thorax bei Patienten mit unklarem akutem Thoraxschmerz. Bisher wurden Untersuchungen bei Patienten mit akutem Thoraxschmerz in Spiraltechnik ohne EKG-Triggerung durchgefuehrt. Das ''Chest-pain-'' oder ''Triple-rule-out''-Protokoll ist ein neues EKG-getriggertes Untersuchungsprotokoll des gesamten Thorax. Im vorliegenden Artikel werden verschiedene Techniken zur CT-Untersuchung von Patienten mit akutem Thoraxschmerz vorgestellt und besprochen. Mit der richtigen Untersuchungstechnik muss die Strahlenexposition fuer ein EKG-getriggertes Untersuchungsprotokoll nicht hoeher sein als eine Standarduntersuchung ohne EKG. Mit einem EKG-getriggerten Untersuchungsprotokoll laesst sich die Aorta in Hinblick auf

  12. Evaluation of body simulator for chest and abdomen in digital X-ray equipment

    International Nuclear Information System (INIS)

    Soares, Sidney S.; Cardoso, Gabriela P.; Oliveira, Giovanni Antônio P.; Batista, Adriana S.M.; Pereira, Esther Lorrayne M.

    2017-01-01

    The use of body simulators to control the quality of X-ray images is a practice that guarantees the control of essential parameters for diagnosis by the technique. The evolution of the equipment, between the analogue, digital computerized radiology (CR) and direct radiography (DR), requires evaluation of the equivalence in grayscale, of simulators, for an adjustment according to the specific technology of obtaining the image. In this sense, the present work presents the evaluation of a body simulator with regard to the representation of mean values of signal, noise and contrast obtained in chest radiographs and panoramic of the abdomen. For the thorax the cardiac region was considered as simulation target and for the abdomen simulation of the liver and small intestine. We used a retrospective study of images obtained with X-ray equipment - CR system, in which the images were studied using the ImageJ program, generating a data catalog. These were subsequently compared with those obtained experimentally using gel filled polymer body simulator. For the validation of the simulator, it was observed the gel equivalence of filling of the polymer box required to reach the image parameters of the cataloged radiographs. The results are discussed as to the physical principles of radiation interaction with biological and equivalent tissues

  13. Nuclear Scan Strategy and Outcomes in Chest Pain Patients Value of Stress Testing with Dipyridamole or Adenosine

    International Nuclear Information System (INIS)

    Conti, Alberto; Mariannini, Yuri; Canuti, Erica; Petrova, Tetyana; Innocenti, Francesca; Zanobetti, Maurizio; Gallini, Chiara; Costanzo, Egidio

    2014-01-01

    To update the prognostic value of scan strategy with pharmacological stress agent in chest pain (CP) patients presenting with normal electrocardiography (ECG) and troponin. Two consecutive nonrandomized series of patients with CP and negative first-line workup inclusive of serial ECG, serial troponin, and echocardiography underwent myocardial perfusion imaging single photon emission computed tomography (SPECT) in the emergency department. Of 170 patients enrolled, 52 patients underwent dipyridamole-SPECT and 118 adenosine-SPECT. Patients with perfusion defects underwent angiography, whereas the remaining patients were discharged and followed-up. Primary endpoint was the composite of nonfatal myocardial infarction, unstable angina, revascularization, and cardiovascular death at follow-up or the presence of coronary stenosis > 50% at angiography. At multivariate analysis, the presence of perfusion defects or hypertension was independent predictor of the primary endpoint. Sensitivity and negative predictive value were higher in patients subjected to adenosine-SPECT (95% and 99%, respectively) versus dipyridamole-SPECT (56% and 89%; yield 70% and 11%, respectively; P < 0.03). Of note, sensitivity, negative, and positive predictive values were high in patients with hypertension (100%, 93%, and 60%, respectively) or nonischemic echocardiography alterations (100%, 100%, and 100%, respectively). In CP patients, presenting with normal ECG and troponin, adenosine-SPECT adds incremental prognostic values to dipyridamole-SPECT. Costly scan strategy is more appropriate and avoids unnecessary angiograms in patients with hypertension or nonischemic echocardiography alterations

  14. Discrete-Slots Models of Visual Working-Memory Response Times

    Science.gov (United States)

    Donkin, Christopher; Nosofsky, Robert M.; Gold, Jason M.; Shiffrin, Richard M.

    2014-01-01

    Much recent research has aimed to establish whether visual working memory (WM) is better characterized by a limited number of discrete all-or-none slots or by a continuous sharing of memory resources. To date, however, researchers have not considered the response-time (RT) predictions of discrete-slots versus shared-resources models. To complement the past research in this field, we formalize a family of mixed-state, discrete-slots models for explaining choice and RTs in tasks of visual WM change detection. In the tasks under investigation, a small set of visual items is presented, followed by a test item in 1 of the studied positions for which a change judgment must be made. According to the models, if the studied item in that position is retained in 1 of the discrete slots, then a memory-based evidence-accumulation process determines the choice and the RT; if the studied item in that position is missing, then a guessing-based accumulation process operates. Observed RT distributions are therefore theorized to arise as probabilistic mixtures of the memory-based and guessing distributions. We formalize an analogous set of continuous shared-resources models. The model classes are tested on individual subjects with both qualitative contrasts and quantitative fits to RT-distribution data. The discrete-slots models provide much better qualitative and quantitative accounts of the RT and choice data than do the shared-resources models, although there is some evidence for “slots plus resources” when memory set size is very small. PMID:24015956

  15. Diagnostic modalities x-ray and CT chest differ in the management of thoracic injury

    Directory of Open Access Journals (Sweden)

    D Chapagain

    2015-06-01

    Full Text Available Objective: To observe difference in the management of blunt trauma to the chest on the basis of conventional xray and computerised tomography of the chest. Methods: This prospective study was conducted between December 2011 to October 2012 in COMS in Bharatpur,a tertiary referral centre in central Nepal . Clinically stable thoracic injury patients were first evaluated with chest x-ray and the management on this basis was recorded. The findings of the CT chest were assessed and the type of management on the basis of CT was also recorded. Outcome was assessed in terms of mortality, morbidity, hospital and ICU stay with respect to the management on the basis of chest x-ray and CT scan. Results: Of the 129 patients, 74.4% were male and 25.6% were female with the patients ranging in age from 7 to 87 years (mean = 40.41 years. The most common mechanism of trauma to the chest was as a result of a motor vehicle accident (69.8%, followed by fall injury (20.2%. X-ray chest diagnosed rib fracture in 62%, haemothorax in 37%, pneumothorax in 27%, lung contusion in 10% and haemopneumothorax in 21% patients. Similarly CT chest diagnosed rib fracture in 86%, haemothorax in 54%, pneumothorax in 36%, lung contusion in 30% and haemopneumothorax in 30% patients. Mean hospital stay was 9.5 days in the group of patients having management on the basis of x-ray chest relative to mean stay of 10.2 days in the CT- chest group. In the management on the basis of xray group, there was a mean ICU stay of 2.8days compared to mean stays of 3.2 days in CT chest group. Conclusion: Though CT scan of the chest is more informative and differs the management of the blunt chest trauma, one should not forget to advise the cost effective, easily available and initial guiding agent, xray chest for early management of the chest injury patient. DOI: http://dx.doi.org/10.3126/jcmsn.v10i1.12764 Journal of College of Medical Sciences-Nepal, 2014, Vol.10(1; 22-31

  16. Diagnostic value of chest radiographs in bedridden patients suspected of having pneumonia.

    Science.gov (United States)

    Esayag, Yaacov; Nikitin, Irina; Bar-Ziv, Jacob; Cytter, Ruth; Hadas-Halpern, Irith; Zalut, Todd; Yinnon, Amos M

    2010-01-01

    To assess the diagnostic value of the chest radiograph for the diagnosis of pneumonia in bedridden patients, using non-contrast-enhanced high-resolution chest computed tomography (CT) as the gold standard. We prospectively evaluated bedridden patients hospitalized with moderate to high clinical probability of pneumonia. Chest radiographs were interpreted in a blinded fashion by 3 observers and classified as definite, normal, or uncertain for pneumonia. Chest CT was obtained within 12 hours of chest radiograph. We applied Bayesian analysis to assess the accuracy of chest radiograph in the diagnosis of pneumonia. In a 5-month period, 58 patients were evaluated, 31 (53%) were female. Their chest radiographs were interpreted as negative, uncertain, or positive for pneumonia in 31 (53%), 15 (26%), and 12 (21%) patients, respectively, while CT confirmed pneumonia in 11 (35%), 10 (67%), and in 10 (83%). The sensitivity of the chest radiograph to diagnose pneumonia was 65%, the specificity was 93%, the positive and negative predictive values were, respectively, 83% and 65%, while the overall accuracy was 69% (95% confidence interval, 50%-79%). In bedridden patients with suspected pneumonia, a normal chest radiograph does not rule out the diagnosis, hence, a chest CT scan might provide valuable diagnostic information. Copyright 2010 Elsevier Inc. All rights reserved.

  17. Comparison of image quality and effective dose by additional filtration on digital chest tomosynthesis

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Kye Sun [Dept. of Dignostic Radiology, Samsung Medical Center, Seoul (Korea, Republic of); Kim, Sung Chul [Dept. of Radiological Science, Gachon University, Sungnam (Korea, Republic of)

    2015-12-15

    The purpose of this study is to suggest proper additional filtration by comparisons patient dose and image quality among additional filters in digital chest tomosynthesis (DTS). We measured the effective dose, dose area product (DAP) by changing thickness of Cu, Al and Ni filter to compare image quality by CD curve and SNR, CNR. Cu, Al and Ni exposure dose were similar thickness 0.3 mm, 3 mm and 0.3 mm respectively. The exposure dose using filter was decreased average about 33.1% than non filter. The DAP value of 0.3 mm Ni were decreased 72.9% compared to non filter and the lowest dose among 3 filter. The effective dose of 0.3 mm Ni were decreased 48% compared to 0.102 mSv effective dose of non filter. At the result of comparison of image quality through CD curve there were similar aspect graph among Cu, Al and Ni. SNR was decreased average 19.07%, CNR was average decreased 18.17% using 3 filters. In conclusion, Ni filtration was considered to be most suitable when considered comprehensive thickness, character, sort of filter, dose reduction and image quality evaluation in DTS.

  18. Slotted Waveguide and Antenna Study for HPM and RF Applications

    Science.gov (United States)

    2017-07-25

    conventional rectangular slot arrays. The proposed CSR-SW A provided higher peak gain than the conventional lambda /4 rectangu lar slot array and a...more compact s ize (55% size reductjon) compared to the lambda /2 rectangular slot array. Both the theoretic and measured results suggest a return loss...in terms of scattering matrix S is: (I-8) With the scattering matrix S, we can compute the scattering field at any position. The resonance of SCCR

  19. Automated lung volumetry from routine thoracic CT scans: how reliable is the result?

    Science.gov (United States)

    Haas, Matthias; Hamm, Bernd; Niehues, Stefan M

    2014-05-01

    Today, lung volumes can be easily calculated from chest computed tomography (CT) scans. Modern postprocessing workstations allow automated volume measurement of data sets acquired. However, there are challenges in the use of lung volume as an indicator of pulmonary disease when it is obtained from routine CT. Intra-individual variation and methodologic aspects have to be considered. Our goal was to assess the reliability of volumetric measurements in routine CT lung scans. Forty adult cancer patients whose lungs were unaffected by the disease underwent routine chest CT scans in 3-month intervals, resulting in a total number of 302 chest CT scans. Lung volume was calculated by automatic volumetry software. On average of 7.2 CT scans were successfully evaluable per patient (range 2-15). Intra-individual changes were assessed. In the set of patients investigated, lung volume was approximately normally distributed, with a mean of 5283 cm(3) (standard deviation = 947 cm(3), skewness = -0.34, and curtosis = 0.16). Between different scans in one and the same patient the median intra-individual standard deviation in lung volume was 853 cm(3) (16% of the mean lung volume). Automatic lung segmentation of routine chest CT scans allows a technically stable estimation of lung volume. However, substantial intra-individual variations have to be considered. A median intra-individual deviation of 16% in lung volume between different routine scans was found. Copyright © 2014 AUR. Published by Elsevier Inc. All rights reserved.

  20. Geothermal reservoirs. Position of slotted section of the tube casing

    International Nuclear Information System (INIS)

    Carotenuto, A.; Vanoli, L.; Casarosa, C.

    1999-01-01

    In the present work the authors have verified the influence of the position of slotted section casing on heat rate drawn by plants for exploitation of geothermal reservoirs that use heat exchangers placed at the bottom of the well (DHE). This study have been done modelling numerically the aquifer, by means of finite element method, evaluating the heat rate drawn by the heat exchanger at different position of the slotted section of the tube casing. Numerical calculations have allowed to show the influence of the main characteristics of the aquifer and of the main characteristics of the aquifer and of the plant on design of the slotted section of the tube casing. In particular, the authors have studied the influence of i) equivalent conductivity and permeability of the aquifer, ii) mass flow rate and the inlet and outlet aquifer temperature difference in the well, iii) the ratio between the length of the slotted section and the thickness of the geothermal layer, varying the position of the slotted section of the tube casing in the aquifer [it

  1. Vertically Polarized Omnidirectional Printed Slot Loop AntennaPrinted Slot Loop Antenna (invited)

    DEFF Research Database (Denmark)

    Kammersgaard, Nikolaj Peter Iversen; Kvist, Søren Helstrup; Thaysen, Jesper

    2015-01-01

    and in-phase fields in the slot in order to obtain an omnidirectional radiation pattern. The antenna is designed for the 2.45 GHz Industrial, Scientific and Medical band. Applications of the antenna are many. One is for on-body applications since it is ideal for launching a creeping waves due...

  2. Vision-Based Parking-Slot Detection: A Benchmark and A Learning-Based Approach

    Directory of Open Access Journals (Sweden)

    Lin Zhang

    2018-03-01

    Full Text Available Recent years have witnessed a growing interest in developing automatic parking systems in the field of intelligent vehicles. However, how to effectively and efficiently locating parking-slots using a vision-based system is still an unresolved issue. Even more seriously, there is no publicly available labeled benchmark dataset for tuning and testing parking-slot detection algorithms. In this paper, we attempt to fill the above-mentioned research gaps to some extent and our contributions are twofold. Firstly, to facilitate the study of vision-based parking-slot detection, a large-scale parking-slot image database is established. This database comprises 8600 surround-view images collected from typical indoor and outdoor parking sites. For each image in this database, the marking-points and parking-slots are carefully labeled. Such a database can serve as a benchmark to design and validate parking-slot detection algorithms. Secondly, a learning-based parking-slot detection approach, namely P S D L , is proposed. Using P S D L , given a surround-view image, the marking-points will be detected first and then the valid parking-slots can be inferred. The efficacy and efficiency of P S D L have been corroborated on our database. It is expected that P S D L can serve as a baseline when the other researchers develop more sophisticated methods.

  3. Dimensional accuracy of digital dental models from cone-beam computed tomography scans of alginate impressions according to time elapsed after the impressions.

    Science.gov (United States)

    Lee, Sang-Mi; Hou, Yanan; Cho, Jin-Hyoung; Hwang, Hyeon-Shik

    2016-02-01

    Our objective was to investigate the dimensional accuracy of digital dental models obtained from the cone-beam computed tomography (CBCT) scans of alginate impressions according to the time elapsed after the impressions were taken. Alginate impressions were obtained of 20 adults using 2 alginate materials: Alginoplast (Heraeus Kulzer, Hanau, Germany) and Cavex (Cavex Holland, Haarlem, The Netherlands). These impressions were stored in wet conditions and scanned by CBCT immediately after impression-taking and after storage times of 12, 24, 36, 48, 60, and 72 hours. After reconstruction of the 3-dimensional digital dental models, the models were measured, and the data were analyzed to determine the dimensional changes according to the time elapsed. The changes within the measurement error were regarded as clinically acceptable in this study. The measurement errors ranged from 0.27 to 0.29 mm in the digital dental models obtained from CBCT scans of the alginate impressions. All measurements showed decreasing accuracy with greater elapsed time after the impressions were taken. Changes above the measurement error occurred at 24 and 36 hours for Cavex and Alginoplast, respectively. Digital dental models can be obtained from CBCT scans of alginate impressions without sending them to a remote laboratory. However, the impressions need to be scanned within 24 hours; otherwise, dimensional changes lead to errors that exceed the error of measurement. Copyright © 2016 American Association of Orthodontists. Published by Elsevier Inc. All rights reserved.

  4. Comparison of Digital Tomosynthesis and Chest Radiography for the Detection of Noncalcified Pulmonary and Hilar Lesions.

    Science.gov (United States)

    Galea, Angela; Adlan, Tarig; Gay, David; Roobottom, Carl; Dubbins, Paul; Riordan, Richard

    2015-09-01

    The aim of this study was to compare the sensitivity and specificity of chest digital tomosynthesis (DTS) with chest radiography (CXR) for the detection of noncalcified pulmonary nodules and hilar lesions using computed tomography (CT) as the reference standard. A total of 78 patients with suspected noncalcified pulmonary lesions on CXR were included in the study. Two radiologists, blinded to the history and CT, analyzed the CXR and the DTS images (separately), whereas a third radiologist analyzed the CXR and DTS images together. Noncalcified intrapulmonary nodules and hilar lesions were recorded for analysis. The interobserver agreement for CXR and DTS was assessed, and the time taken to report the images was recorded. A total of 202 lesions were recorded in 78 patients. There were 111 true lesions confirmed on CT in 53 patients; in 25 patients subsequent CT excluded a lesion. The overall sensitivity was 32% for CXR and 49% for DTS. This improved to 54% when the posteroanterior CXR and DTS were reviewed together (CXR-DTS). The overall specificities for CXR, DTS, and CXR-DTS were 49%, 96%, and 98%, respectively. There were 56 suspected hilar lesions with subgroup sensitivities of 76% for CXR, 65% for DTS, and 76% for CXR-DTS. The specificity for hilar lesions was 59%, 92%, and 97% for CXR, DTS, and CXR-DTS, respectively. DTS significantly improves the detectability of noncalcified nodules when compared with and when used in combination with CXR. The specificity and interobserver agreement of DTS in the diagnosis of suspected noncalcified pulmonary nodules and hilar lesions are significantly better than those of CXR and approaches those of CT.

  5. Modeling of Slot Waveguide Sensors Based on Polymeric Materials

    Science.gov (United States)

    Bettotti, Paolo; Pitanti, Alessandro; Rigo, Eveline; De Leonardis, Francesco; Passaro, Vittorio M. N.; Pavesi, Lorenzo

    2011-01-01

    Slot waveguides are very promising for optical sensing applications because of their peculiar spatial mode profile. In this paper we have carried out a detailed analysis of mode confinement properties in slot waveguides realized in very low refractive index materials. We show that the sensitivity of a slot waveguide is not directly related to the refractive index contrast of high and low materials forming the waveguide. Thus, a careful design of the structures allows the realization of high sensitivity devices even in very low refractive index materials (e.g., polymers) to be achieved. Advantages of low index dielectrics in terms of cost, functionalization and ease of fabrication are discussed while keeping both CMOS compatibility and integrable design schemes. Finally, applications of low index slot waveguides as substitute of bulky fiber capillary sensors or in ring resonator architectures are addressed. Theoretical results of this work are relevant to well established polymer technologies. PMID:22164020

  6. Modeling of Slot Waveguide Sensors Based on Polymeric Materials

    Directory of Open Access Journals (Sweden)

    Lorenzo Pavesi

    2011-07-01

    Full Text Available Slot waveguides are very promising for optical sensing applications because of their peculiar spatial mode profile. In this paper we have carried out a detailed analysis of mode confinement properties in slot waveguides realized in very low refractive index materials. We show that the sensitivity of a slot waveguide is not directly related to the refractive index contrast of high and low materials forming the waveguide. Thus, a careful design of the structures allows the realization of high sensitivity devices even in very low refractive index materials (e.g., polymers to be achieved. Advantages of low index dielectrics in terms of cost, functionalization and ease of fabrication are discussed while keeping both CMOS compatibility and integrable design schemes. Finally, applications of low index slot waveguides as substitute of bulky fiber capillary sensors or in ring resonator architectures are addressed. Theoretical results of this work are relevant to well established polymer technologies.

  7. Investigation of slot discharge on a 239 MVA hydro generator stator winding

    Energy Technology Data Exchange (ETDEWEB)

    Li, S.; Hong, W. [BC Hydro and Power Authority, Vancouver, BC (Canada)

    2009-07-01

    This paper discussed a slot discharge investigation conducted on a 239 MVA generator stator winding. The generator in which the winding was located had experienced core split distortion, stator winding phase-to-phase failures, winding failures during Hipot testing, and high partial discharge (PD) activity. The results of on-line PD testing data were evaluated. The stator winding was subjected to visual inspections, bar dissections, and failure mechanism analyses. Eleven winding bars were removed from the stator slots in order to assess groundwall insulation conditions and identify the cause of the slot discharge activity. It was determined that the root cause of the slot discharge was a loose, non-uniform bar in the slot. The vibrating bar caused the semi-conductive coating to wear out and degraded the armour tape. Results of the study demonstrated the importance of on-line PD monitoring for detecting slot PD activity. 4 refs., 3 tabs., 11 figs.

  8. Collaborative en-route and slot allocation algorithm based on fuzzy comprehensive evaluation

    Science.gov (United States)

    Yang, Shangwen; Guo, Baohua; Xiao, Xuefei; Gao, Haichao

    2018-01-01

    To allocate the en-routes and slots to the flights with collaborative decision making, a collaborative en-route and slot allocation algorithm based on fuzzy comprehensive evaluation was proposed. Evaluation indexes include flight delay costs, delay time and the number of turning points. Analytic hierarchy process is applied to determining index weights. Remark set for current two flights not yet obtained the en-route and slot in flight schedule is established. Then, fuzzy comprehensive evaluation is performed, and the en-route and slot for the current two flights are determined. Continue selecting the flight not yet obtained an en-route and a slot in flight schedule. Perform fuzzy comprehensive evaluation until all flights have obtained the en-routes and slots. MatlabR2007b was applied to numerical test based on the simulated data of a civil en-route. Test results show that, compared with the traditional strategy of first come first service, the algorithm gains better effect. The effectiveness of the algorithm was verified.

  9. Intrathoracic tuberculous lymphadenopathy in children: a guide to chest radiography

    Energy Technology Data Exchange (ETDEWEB)

    George, Anthony; Andronikou, Savvas [Bristol Royal Hospital for Children and the University of Bristol, Department of Paediatric Radiology, Bristol (United Kingdom); Pillay, Tanyia; Zar, Heather J. [University of Cape Town and Medical Research Council Unit on Child and Adolescent Health, Department of Paediatrics and Child Health, Red Cross War Memorial Children' s Hospital, Cape Town (South Africa); Goussard, Pierre [Tygerberg Hospital and the University of Stellenbosch, Department of Paediatrics and Child Health, Cape Town (South Africa)

    2017-09-15

    Making the diagnosis of pulmonary tuberculosis in children can be difficult because microbiological confirmation is not often achieved. Diagnosis is therefore often based on clinical features in combination with chest radiograph findings. Chest radiographs can demonstrate lymphadenopathy of the hilar and para-tracheal regions on the anteroposterior view, and subcarinal lymphadenopathy on the lateral view. However poor interobserver agreement has been reported for radiologist and clinician assessment of lymphadenopathy. This might reflect the lack of standardised imaging criteria for diagnosis as well as radiologists' objectives for achieving sensitivity rather than specificity. In this paper the authors provide a pictorial aid of chest radiographs in children with culture-confirmed tuberculosis to help clinicians identify lymph node enlargement in primary pulmonary tuberculosis. This collection of images comprises chest radiographs accompanied by schematics and either CT or MRI scan confirmation of pathological lymph node enlargement at the positions commonly affected in tuberculosis. (orig.)

  10. Study of 67Ga scan in sarcoidosis

    International Nuclear Information System (INIS)

    Han Lijun; Qu Wanyin; Liu Xiuqin

    1997-01-01

    Gallium scan and serum angiotensin-converting enzyme assay (SACE) were compared in patients with sarcoidosis. The examination of 67 Ga scan, SACE determination, pulmonary function test, chest CT and chest X-ray in 24 cases with sarcoidosis were studied. The results revealed that 4 of 24 cases had obviously high uptake of 67 Ga exceeding hepatic activity (3+) in clinical active stage, 3 patients had resembling the Greek letter lambda, symmetrically located in bilateral hilar lymph nodes, and among them two had an uptake of 67 Ga in the bilateral lacrimal and parotid gland simulating 'Panda Face'. 8 of 20 cases with inactive sarcoidosis had an abnormal 67 Ga scan (1+). In those patients with normal SACE level but increased uptake of 67 Ga, active stage of disease was demonstrated and steroid therapy was indicated. Gallium scan is a valuable method for the staging of its activity and evaluation of the therapeutic effect in the follow-up patients with sarcoidosis

  11. Pulmonary ventilation/perfusion scan

    Science.gov (United States)

    ... to stop eating (fast), be on a special diet, or take any medicines before the test. A chest x-ray is usually done before or after a ventilation and perfusion scan. You wear a hospital gown or comfortable clothing that does not have ...

  12. What did we learn from two decades of chest computed tomography in cystic fibrosis?

    International Nuclear Information System (INIS)

    Tiddens, Harm A.W.M.; Rosenow, Tim

    2014-01-01

    Despite our current treatment, many cystic fibrosis (CF) patients still show progressive bronchiectasis and small airways disease. Adequate detection and monitoring of progression of these structural abnormalities is needed to personalize treatment to the severity of CF lung disease of the patient. Chest computed tomography (CT) is the gold standard to diagnose and monitor bronchiectasis. Many studies have been done to validate the role of chest CT in CF and to improve the protocols. From these studies it became clear that for correct interpretation of the severity of bronchiectasis and small airways disease standardization of lung volume for the inspiratory and expiratory CT scan acquisition is needed. The risk related to the radiation exposure of a chest CT scan every second year is considered low. Automated and quantitative image analysis systems are developed to improve the reliability and sensitivity of assessments of structural lung changes in CF, particularly in early life. In this paper an overview is given of the lessons learned from two decades of monitoring CF lung disease using chest CT. (orig.)

  13. What did we learn from two decades of chest computed tomography in cystic fibrosis?

    Energy Technology Data Exchange (ETDEWEB)

    Tiddens, Harm A.W.M. [Erasmus MC-Sophia Children' s Hospital, Department of Pediatric Pulmonology, Rotterdam (Netherlands); Erasmus MC, Department of Radiology, Rotterdam (Netherlands); Rosenow, Tim [Erasmus MC-Sophia Children' s Hospital, Department of Pediatric Pulmonology, Rotterdam (Netherlands); The University of Western Australia, Telethon Institute for Child Health Research, Perth (Australia); The University of Western Australia, School of Paediatrics and Child Health Research, Perth (Australia)

    2014-12-15

    Despite our current treatment, many cystic fibrosis (CF) patients still show progressive bronchiectasis and small airways disease. Adequate detection and monitoring of progression of these structural abnormalities is needed to personalize treatment to the severity of CF lung disease of the patient. Chest computed tomography (CT) is the gold standard to diagnose and monitor bronchiectasis. Many studies have been done to validate the role of chest CT in CF and to improve the protocols. From these studies it became clear that for correct interpretation of the severity of bronchiectasis and small airways disease standardization of lung volume for the inspiratory and expiratory CT scan acquisition is needed. The risk related to the radiation exposure of a chest CT scan every second year is considered low. Automated and quantitative image analysis systems are developed to improve the reliability and sensitivity of assessments of structural lung changes in CF, particularly in early life. In this paper an overview is given of the lessons learned from two decades of monitoring CF lung disease using chest CT. (orig.)

  14. A multi-slot surface coil for MRI of dual-rat imaging at 4 T

    International Nuclear Information System (INIS)

    Solis, S E; Rodriguez, A O; Wang, R; Tomasi, D

    2011-01-01

    A slotted surface coil inspired by the hole-and-slot cavity magnetron was developed for magnetic resonance imaging of obese rats at 4 T. Full-wave analysis of the magnetic field was carried out at 170 MHz for both the slotted and circular-shaped coils. The noise figure values of two coils were investigated via the numerical calculation of the quality factors. Fat simulated phantoms to mimic overweight rats were included in the analysis with weights ranging from 300 to 900 g. The noise figures were 1.2 dB for the slotted coil and 2.4 dB for the circular coil when loaded with 600 g of simulated phantom. A slotted surface coil with eight circular slots and a circular coil with similar dimensions were built and operated in the transceiver mode, and their performances were experimentally compared. The imaging tests in phantoms demonstrated that the slotted surface coil has a deeper RF-sensitivity and better field uniformity than the single-loop RF-coil. High quality images of two overweight Zucker rats were acquired simultaneously with the slotted surface coil using standard spin-echo pulse sequences. Experimental results showed that the slotted surface coil outperformed the circular coil for imaging considerably overweight rats. Thus, the slotted surface coil can be a good tool for MRI experiments in rats on a human whole-body 4 T scanner.

  15. A multi-slot surface coil for MRI of dual-rat imaging at 4T

    Energy Technology Data Exchange (ETDEWEB)

    Solis, S.E.; Tomasi, D.; Solis, S.E.; Wang, R.; Tomasi, D.; Rodriguez, A.O.

    2011-07-01

    A slotted surface coil inspired by the hole-and-slot cavity magnetron was developed for magnetic resonance imaging of obese rats at 4 T. Full-wave analysis of the magnetic field was carried out at 170 MHz for both the slotted and circular-shaped coils. The noise figure values of two coils were investigated via the numerical calculation of the quality factors. Fat simulated phantoms to mimic overweight rats were included in the analysis with weights ranging from 300 to 900 g. The noise figures were 1.2 dB for the slotted coil and 2.4 dB for the circular coil when loaded with 600 g of simulated phantom. A slotted surface coil with eight circular slots and a circular coil with similar dimensions were built and operated in the transceiver mode, and their performances were experimentally compared. The imaging tests in phantoms demonstrated that the slotted surface coil has a deeper RF-sensitivity and better field uniformity than the single-loop RF-coil. High quality images of two overweight Zucker rats were acquired simultaneously with the slotted surface coil using standard spin-echo pulse sequences. Experimental results showed that the slotted surface coil outperformed the circular coil for imaging considerably overweight rats. Thus, the slotted surface coil can be a good tool for MRI experiments in rats on a human whole-body 4 T scanner.

  16. A multi-slot surface coil for MRI of dual-rat imaging at 4 T

    Energy Technology Data Exchange (ETDEWEB)

    Solis, S E; Rodriguez, A O [Departamento de Ingenieria Electrica, Universidad Autonoma Metropolitana Iztapalapa, Mexico, DF 09340 (Mexico); Wang, R; Tomasi, D, E-mail: arog@xanum.uam.mx [Medical Department, Brookhaven National Laboratory, Upton, NY 11973 (United States)

    2011-06-21

    A slotted surface coil inspired by the hole-and-slot cavity magnetron was developed for magnetic resonance imaging of obese rats at 4 T. Full-wave analysis of the magnetic field was carried out at 170 MHz for both the slotted and circular-shaped coils. The noise figure values of two coils were investigated via the numerical calculation of the quality factors. Fat simulated phantoms to mimic overweight rats were included in the analysis with weights ranging from 300 to 900 g. The noise figures were 1.2 dB for the slotted coil and 2.4 dB for the circular coil when loaded with 600 g of simulated phantom. A slotted surface coil with eight circular slots and a circular coil with similar dimensions were built and operated in the transceiver mode, and their performances were experimentally compared. The imaging tests in phantoms demonstrated that the slotted surface coil has a deeper RF-sensitivity and better field uniformity than the single-loop RF-coil. High quality images of two overweight Zucker rats were acquired simultaneously with the slotted surface coil using standard spin-echo pulse sequences. Experimental results showed that the slotted surface coil outperformed the circular coil for imaging considerably overweight rats. Thus, the slotted surface coil can be a good tool for MRI experiments in rats on a human whole-body 4 T scanner.

  17. Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans

    International Nuclear Information System (INIS)

    Tappouni, Rafel; Mathers, Bradley

    2012-01-01

    Objectives. To compare the effectiveness of the bismuth breast shield and partial CT scan in reducing entrance skin dose and to evaluate the effect of the breast shield on image quality (IQ). Methods. Nanodots were placed on an adult anthropomorphic phantom. Standard chest CT, CT with shield, and partial CT were performed. Nanodot readings and effective doses were recorded. 50 patients with chest CTs obtained both with and without breast shields were reviewed. IQ was evaluated by two radiologists and by measuring Hounsfield units (HUs) and standard deviation (SD) of HU in anterior subcutaneous region. Results. Breast shield and the partial CT scans reduced radiation to the anterior chest by 38% and 16%, respectively. Partial CT increased dose to the posterior chest by 37% and effective dose by 8%. Change in IQ in shield CT was observed in the anterior chest wall. Significant change in IQ was observed in 5/50 cases. The shield caused an increase of 20 HU (P = 0.021) and a 1.86 reduction in SD of HU (P = 0.027) in the anterior compared to posterior subcutaneous regions. Summary. Bismuth breast shield is more effective than the partial CT in reducing entrance skin dose while maintaining image quality

  18. Influence of scanning system and dentist's level of training in the accuracy of digital impressions

    Science.gov (United States)

    Hategan, Simona; Gabor, Alin; Zaharia, Cristian; Sinescu, Cosmin; Negrutiu, Meda Lavinia; Jivanescu, Anca

    2016-03-01

    Background: The principal aim of our study was to evaluate digital impressions, taken with spray powder and powderfree scan systems, in order to determine the influence of the dentist's commitment to training as a critical factor regarding quality. Material and method: Two digital intraoral impression systems from the same manufacture (Sirona) : Apollo DI and CEREC Omnicam, were used to scan 16 crown preparations on teeth on a typodont maxillary model. Because an Apollo Di intraoral camera is a powder system, an adhesive was applied before using the powder spray. Three groups were used to scan the crown preparations in order to determine coating thickness homogeneity. One group consisted of senior year dental students, a second consisted of prosthodontics residents, and the third consisted of prosthodontics specialists. The same procedure was applied with a CEREC Omnicam intraoral camera, which is a powder-free system. By using the two systems software parameters we were able to determine the scanning precision. Results: Homogeneity scores for Apollo Di regarding the spray layer was significantly thinner for all dental surfaces in the first group, while the second group had thinner coatings for buccal and distal surfaces. For the third group, the crown preparations were coated more homogeneously than the first two groups. The powder-free system CEREC Omnicam can, to a degree, mask the lack of experience in direct optical impressions by avoiding the poor quality coating, which can lead to defective marginal adaptation of definitive restoration. Conclusions: The dentist's lack of experience can be mitigated, and partially avoided, by using powder-free systems. At the same time, the dentist can give more time towards learning how to integrate computerized fabricated restoration into the practice. The commitment to training is a critical factor in the successful integration of the technology. In addition, scanning marginal preparation details needs time in order to

  19. Lateral topography for reducing effective dose in low-dose chest CT.

    Science.gov (United States)

    Bang, Dong-Ho; Lim, Daekeon; Hwang, Wi-Sub; Park, Seong-Hoon; Jeong, Ok-man; Kang, Kyung Wook; Kang, Hohyung

    2013-06-01

    The purposes of this study were to assess radiation exposure during low-dose chest CT by using lateral topography and to compare the lateral topographic findings with findings obtained with anteroposterior topography alone and anteroposterior and lateral topography combined. From November 2011 to February 2012, 210 male subjects were enrolled in the study. Age, weight, and height of the men were recorded. All subjects were placed into one of three subgroups based on the type of topographic image obtained: anteroposterior topography, lateral topography, and both anteroposterior and lateral topography. Imaging was performed with a 128-MDCT scanner. CT, except for topography, was the same for all subjects. A radiologist analyzed each image, recorded scan length, checked for any insufficiencies in the FOV, and calculated the effective radiation dose. One-way analysis of variance and multiple comparisons were used to compare the effective radiation exposure and scan length between groups. The mean scan length in the anteroposterior topography group was significantly greater than that of the lateral topography group and the combined anteroposterior and lateral topography group (p topography group (0.735 ± 0.033 mSv) was significantly lower than that for the anteroposterior topography group (0.763 ± 0.038 mSv) and the combined anteroposterior and lateral topography group (0.773 ± 0.038) (p < 0.001). Lateral topographic low-dose CT was associated with a lower effective radiation dose and scan length than either anteroposterior topographic low-dose chest CT or low-dose chest CT with both anteroposterior and lateral topograms.

  20. Energy characteristics of the double slot in the narrow wall of a rectangular waveguide

    OpenAIRE

    Martynenko, S. A.

    2005-01-01

    Based on approximation of the half-wave field distribution in the slots, an expression is derived for internal mutual conductance of closely-spaced slots, which form a double inclined slot in the narrow wall of a rectangular waveguide. The narrow wall has cut-outs reaching the broad wall. With the use of the method of induced magnetomotive forces, a mathematical model is devised for calculating the energy characteristics of the double slot. The impact of angle of inclination of the slots, dim...

  1. Initial application of digital tomosynthesis for detection of pulmonary nodules

    International Nuclear Information System (INIS)

    Sun Zhiyuan; Su Hong; Zhao Yane; Ju Bing; Chang Shuanghui; Hu Qiuju; Lu Guangming

    2010-01-01

    Objective: To discuss the value of digital tomosynthesis for detection of pulmonary nodules. Methods: Thirty patients suspected of having pulmonary nodules underwent chest radiography, digital tomosynthesis and CT examination. Above image data were transferred to postprocessing work station and were reviewed by 2 radiologists with 3 years of chest-radiology diagnosis experience in a double-blind method. The number, location and size of nodules were recorded. Then, 2 radiologists reviewed the all images once more, and discuss in consensus. The sensitivities of chest radiography and digital tomosynthesis for detection of pulmonary nodules were respectively calculated according to the CT results. Chi-square test was used for radiography, digital tomosynthesis and CT examination. Results: Of 30 patients, 21 were detected having pulmonary nodules by X-ray radiography and 9 were negative, the total number of 40 nodules was detected, while 89 nodules in 26 patients were detected by digital tomosynthesis, and only 4 patients were negative. CT demonstrated 102 nodules in 27 patients, and 3 patients were negative. Taking CT as 'gold standard', the sensitivities of X-ray radiography and digital tomosynthesis were 27.4%(28/102)and 87.2%(89/102), X 2 =4.35, P<0.05, respectively. Conclusion: Digital tomosynthesis has a high sensitivity for detection of pulmonary nodules compared with X-ray radiography, and could be an excellent and necessary supplementary technique of X-ray radiography. (authors)

  2. Tree-structured vector quantization of CT chest scans: Image quality and diagnostic accuracy

    International Nuclear Information System (INIS)

    Cosman, P.C.; Tseng, C.; Gray, R.M.; Olshen, R.A.; Moses, L.E.; Davidson, H.C.; Bergin, C.J.; Riskin, E.A.

    1993-01-01

    The quality of lossy compressed images is often characterized by signal-to-noise ratios, informal tests of subjective quality, or receiver operating characteristic (ROC) curves that include subjective appraisals of the value of an image for a particular application. The authors believe that for medical applications, lossy compressed images should be judged by a more natural and fundamental aspect of relative image quality: their use in making accurate diagnoses. They apply a lossy compression algorithm to medical images, and quantify the quality of the images by the diagnostic performance of radiologists, as well as by traditional signal-to-noise ratios and subjective ratings. The study is unlike previous studies of the effects of lossy compression in that they consider non-binary detection tasks, simulate actual diagnostic practice instead of using paired tests or confidence rankings, use statistical methods that are more appropriate for non-binary clinical data than are the popular ROC curves, and use low-complexity predictive tree-structured vector quantization for compression rather than DCT-based transform codes combined with entropy coding. Their diagnostic tasks are the identification of nodules (tumors) in the lungs and lymphadenopathy in the mediastinum from computerized tomography (CT) chest scans. For the image modality, compression algorithm, and diagnostic tasks they consider, the original 12 bit per pixel (bpp) CT image can be compressed to between 1 bpp and 2 bpp with no significant changes in diagnostic accuracy

  3. Broadcast Coded Slotted ALOHA

    DEFF Research Database (Denmark)

    Ivanov, Mikhail; Brännström, Frederik; Graell i Amat, Alexandre

    2016-01-01

    We propose an uncoordinated medium access control (MAC) protocol, called all-to-all broadcast coded slotted ALOHA (B-CSA) for reliable all-to-all broadcast with strict latency constraints. In B-CSA, each user acts as both transmitter and receiver in a half-duplex mode. The half-duplex mode gives ...

  4. Results of pressurized-slot measurements in the G-Tunnel underground facility

    International Nuclear Information System (INIS)

    Zimmerman, R.M.; Mann, K.L.; Dodds, D.J.

    1989-01-01

    A rock-mechanics field-testing program is underway at Sandia National Laboratories (SNL) as part of the YMP. SNL has the responsibility for assessing the repository design and performance as well as characterizing the geomechanical behavior of the rock. SNL has conducted field experiments in G-Tunnel in Rainier Mesa at the NTS, where tuffs similar to those at Yucca Mountain, the potential repository site, are found. Later experiments are planned as part of the YMP Exploratory Shaft investigations at Yucca Mountain. Major geomechanical factors in repository developments are determinations of the stress state and the deformability of the rock mass (described by the modulus of deformation). One feature of SNL's rock-mechanics program was the development of a testing program for cutting thin slots in a jointed welded tuff and utilizing flatjacks for pressurizing these thin-slots on a relatively, large scale. Objectives in the pressurized-slot testing in G-Tunnel have been to apply and possibly improve methods for (1) utilizing the flatjack cancellation (FC) method for measuring stresses normal to the slot and (2) measuring the modulus of deformation of the jointed rock surrounding the slot. This paper discusses the results of field measurements in and around a single slot and evaluates potential applications and limitations. 10 refs., 1 fig., 4 tabs

  5. A Community-acquired Lung Abscess Attributable to Streptococcus pneumoniae which Extended Directly into the Chest Wall.

    Science.gov (United States)

    Ko, Yuki; Tobino, Kazunori; Yasuda, Yuichiro; Sueyasu, Takuto; Nishizawa, Saori; Yoshimine, Kouhei; Munechika, Miyuki; Asaji, Mina; Yamaji, Yoshikazu; Tsuruno, Kosuke; Miyajima, Hiroyuki; Mukasa, Yosuke; Ebi, Noriyuki

    We herein report the case of 75-year-old Japanese female with a community-acquired lung abscess attributable to Streptococcus pneumoniae (S. penumoniae) which extended into the chest wall. The patient was admitted to our hospital with a painful mass on the left anterior chest wall. A contrast-enhanced chest computed tomography scan showed a lung abscess in the left upper lobe which extended into the chest wall. Surgical debridement of the chest wall abscess and percutaneous transthoracic tube drainage of the lung abscess were performed. A culture of the drainage specimen yielded S. pneumoniae. The patient showed a remarkable improvement after the initiation of intravenous antibiotic therapy.

  6. A digital transducer and digital microphone using an optical technique

    Science.gov (United States)

    Ghelmansarai, F. A.

    1996-09-01

    A transducer is devised to measure pressure, displacements or angles by optical means. This transducer delivers a digital output without relying on interferometry techniques or analogue-to-digital converters. This device is based on an optical scanner and an optical detector. An inter-digital photoconductive detector (IDPC) is employed that delivers a series of pulses, whose number depends on the scan length. A pre-objective scanning configuration is used that allows for the possibility of a flat image plane. The optical scanner provides scanning of IDPC and the generated scan length is proportional to the measurand.

  7. Image Format Conversion to DICOM and Lookup Table Conversion to Presentation Value of the Japanese Society of Radiological Technology (JSRT) Standard Digital Image Database.

    Science.gov (United States)

    Yanagita, Satoshi; Imahana, Masato; Suwa, Kazuaki; Sugimura, Hitomi; Nishiki, Masayuki

    2016-01-01

    Japanese Society of Radiological Technology (JSRT) standard digital image database contains many useful cases of chest X-ray images, and has been used in many state-of-the-art researches. However, the pixel values of all the images are simply digitized as relative density values by utilizing a scanned film digitizer. As a result, the pixel values are completely different from the standardized display system input value of digital imaging and communications in medicine (DICOM), called presentation value (P-value), which can maintain a visual consistency when observing images using different display luminance. Therefore, we converted all the images from JSRT standard digital image database to DICOM format followed by the conversion of the pixel values to P-value using an original program developed by ourselves. Consequently, JSRT standard digital image database has been modified so that the visual consistency of images is maintained among different luminance displays.

  8. Slot-dimer babinet metamaterials as polarization shapers for terahertz waves

    DEFF Research Database (Denmark)

    Zhukovsky, Sergei; Chigrin, D. N.; Lavrinenko, Andrei

    2013-01-01

    We theoretically study optical properties of free-standing metallic membranes patterned with an array of two-slot elements (dimers) comprising two rectangular slots of different dimensions and orientation. It is shown that these structures feature extraordinary optical transmission with strong...

  9. Mitral valve plasty for mitral regurgitation after blunt chest trauma.

    Science.gov (United States)

    Kumagai, H; Hamanaka, Y; Hirai, S; Mitsui, N; Kobayashi, T

    2001-06-01

    A 21 year-old woman was admitted to our hospital because of chest and back pain after blunt chest trauma. On admission, consciousness was clear and a physical examination showed labored breathing. Her vital signs were stable, but her breathing gradually worsened, and artificial respiration was started. The chest roentgenogram and a subsequent chest computed tomographic scans revealed contusions, hemothorax of the left lung and multiple rib fractures. A transthoracic echocardiography (TTE) revealed normal left ventricular wall motion and mild mitral regurgitation (MR). TTE was carried out repeatedly, and revealed gradually progressive MR and prolapse of the posterior medial leaflet, although there was no congestive heart failure. After her general condition had recovered, surgery was performed. Intraoperative transesophageal echocardiography (TEE) revealed torn chordae at the posterior medial leaflet. The leaflet where the chorda was torn was cut and plicated, and posterior mitral annuloplasty was performed using a prosthetic ring. One month later following discharge, the MR had disappeared on TTE.

  10. Crossed-Slot Cavity-Backed Antenna with Improved Hemispherical Coverage

    DEFF Research Database (Denmark)

    Kim, Oleksiy S.; Breinbjerg, Olav; Østergaard, Allan

    2005-01-01

    The paper presents the results of the investigation of the crossed-slot cavity-backed antenna with the complementary crossed electric dipoles added to compensate the circularly polarized (CP) radiation pattern degradation near the horizon. Dependences of the radiation characteristics...... of the modified crossed-slot cavity-backed antenna on the length, width and height of the crossed electric dipoles are shown. Effects of a finite size ground plane are taken into account due to a full wave electromagnetic analysis software utilized in the parametrical investigations. Simulated and measured...... results for a selected antenna configuration prove that the properly adjusted crossed electric dipoles are able to improve the coverage and CP polarization characteristics of the crossed-slot cavity-backed antenna....

  11. Diagnostic accuracy of oblique chest radiograph for occult pneumothorax: comparison with ultrasonography

    OpenAIRE

    Matsumoto, Shokei; Sekine, Kazuhiko; Funabiki, Tomohiro; Orita, Tomohiko; Shimizu, Masayuki; Hayashida, Kei; Kazamaki, Taku; Suzuki, Tatsuya; Kishikawa, Masanobu; Yamazaki, Motoyasu; Kitano, Mitsuhide

    2016-01-01

    Backgraound An occult pneumothorax is a pneumothorax that is not seen on a supine chest X-ray but is detected by computed tomography scanning. However, critical patients are difficult to transport to the computed tomography suite. We previously reported a method to detect occult pneumothorax using oblique chest radiography (OXR). Several authors have also reported that ultrasonography is an effective technique for detecting occult pneumothorax. The aim of this study was to evaluate the useful...

  12. Feasibility study for image reconstruction in circular digital tomosynthesis (CDTS) from limited-scan angle data based on compressed-sensing theory

    Energy Technology Data Exchange (ETDEWEB)

    Park, Yeonok; Je, Uikyu; Cho, Hyosung, E-mail: hscho1@yonsei.ac.kr; Hong, Daeki; Park, Chulkyu; Cho, Heemoon; Choi, Sungil; Woo, Taeho

    2015-03-21

    In this work, we performed a feasibility study for image reconstruction in a circular digital tomosynthesis (CDTS) from limited-scan angle data based on compressed-sensing (CS) theory. Here, the X-ray source moves along an arc within a limited-scan angle (≤ 180°) on a circular path set perpendicularly to the axial direction during the image acquisition. This geometry, compared to full-angle (360°) scan geometry, allows imaging system to be designed more compactly and gives better tomographic quality than conventional linear digital tomosynthesis (DTS). We implemented an efficient CS-based reconstruction algorithm for the proposed geometry and performed systematic simulations to investigate the image characteristics. We successfully reconstructed CDTS images with incomplete projections acquired at several selected limited-scan angles of 45°, 90°, 135°, and 180° for a given tomographic angle of 80° and evaluated the reconstruction quality. Our simulation results indicate that the proposed method can provide superior tomographic quality for axial view and even for the other views (i.e., sagittal and coronal), as in computed tomography, to conventional DTS. - Highlights: • Image reconstruction is done in circular digital tomosynthesis (CDTS). • The designed geometry allows imaging system to be the better image. • An efficient compressed-sensing (CS)-based reconstruction algorithm is performed. • Proposed method can provide superior tomographic quality for the axial view.

  13. Measurement of the depth of narrow slotted sections in eddy current reference standards

    Science.gov (United States)

    Kim, Young-Joo; Kim, Young-gil; Ahn, Bongyoung; Yoon, Dong-Jin

    2007-02-01

    The dimensions of the slots in eddy current (EC) reference standards are too narrow to be measured by general depth measurement methods such as the optical (laser) or stylus methods. However, measurement of the dimensions of the machined slots is a prerequisite to using the blocks as references. The present paper suggests a measurement method for the slotted section using an ultrasonic test. The width and depth of the slots measured in our study are roughly 0.1 mm and 0.5 mm, respectively. The time of flight (TOF) of the ultrasonic wave was measured precisely. The ultrasonic velocity in the material of the EC reference standard was calculated with the measured values of the TOF and its thickness. Reflected waves from the tip of the slot and the bottom surface of the EC standard were successfully classified. Using this method we have successfully determined the depth of the slotted section.

  14. The transverse diameter of the chest on routine radiographs reliably estimates gestational age and weight in premature infants

    Energy Technology Data Exchange (ETDEWEB)

    Dietz, Kelly R. [University of Minnesota, Department of Radiology, Minneapolis, MN (United States); Zhang, Lei [University of Minnesota, Biostatistical Design and Analysis Center, Minneapolis, MN (United States); Seidel, Frank G. [Lucile Packard Children' s Hospital, Department of Radiology, Stanford, CA (United States)

    2015-08-15

    Prior to digital radiography it was possible for a radiologist to easily estimate the size of a patient on an analog film. Because variable magnification may be applied at the time of processing an image, it is now more difficult to visually estimate an infant's size on the monitor. Since gestational age and weight significantly impact the differential diagnosis of neonatal diseases and determine the expected size of kidneys or appearance of the brain by MRI or US, this information is useful to a pediatric radiologist. Although this information may be present in the electronic medical record, it is frequently not readily available to the pediatric radiologist at the time of image interpretation. To determine if there was a correlation between gestational age and weight of a premature infant with their transverse chest diameter (rib to rib) on admission chest radiographs. This retrospective study was approved by the institutional review board, which waived informed consent. The maximum transverse chest diameter outer rib to outer rib was measured on admission portable chest radiographs of 464 patients admitted to the neonatal intensive care unit (NICU) during the 2010 calendar year. Regression analysis was used to investigate the association between chest diameter and gestational age/birth weight. Quadratic term of chest diameter was used in the regression model. Chest diameter was statistically significantly associated with both gestational age (P < 0.0001) and birth weight (P < 0.0001). An infant's gestational age and birth weight can be reliably estimated by comparing a simple measurement of the transverse chest diameter on digital chest radiograph with the tables and graphs in our study. (orig.)

  15. Slot-type pickup/kicker for AA stochastic cooling

    CERN Multimedia

    CERN PhotoLab

    1979-01-01

    A "slotted transmission line" was used for both pickups and kickers of the cooling systems of the AA. They served for the cooling of the high-density antiproton stack, in momentum and both transverse planes. In the beginning in a single band, 1-2 GHz, later in 2 bands, 2-4 and 4-8 GHz. Here we see the slotted electrodes partly pulled out of the outer casing. See also 7906189, 7906581X, 7896193.

  16. Limiting chest computed tomography in the evaluation of pediatric thoracic trauma.

    Science.gov (United States)

    Golden, Jamie; Isani, Mubina; Bowling, Jordan; Zagory, Jessica; Goodhue, Catherine J; Burke, Rita V; Upperman, Jeffrey S; Gayer, Christopher P

    2016-08-01

    Computed tomography (CT) of the chest (chest CT) is overused in blunt pediatric thoracic trauma. Chest CT adds to the diagnosis of thoracic injury but rarely changes patient management. We sought to identify a subset of blunt pediatric trauma patients who would benefit from a screening chest CT based on their admission chest x-ray (CXR) findings. We hypothesize that limiting chest CT to patients with an abnormal mediastinal silhouette identifies intrathoracic vascular injuries not otherwise seen on CXR. All blunt trauma activations that underwent an admission CXR at our Level 1 pediatric trauma center from 2005 to 2013 were retrospectively reviewed. Patients who had a chest CT were evaluated for added diagnoses and change in management after CT. An admission CXR was performed in 1,035 patients. One hundred thirty-nine patients had a CT, and the diagnosis of intra-thoracic injury was added in 42% of patients. Chest CT significantly increased the diagnosis of contusion or atelectasis (30.3% vs 60.4%; p pneumothorax (7.2% vs 18.7%; p chest CT. Chest CT changed the management of only 4 patients (2.9%). Two patients underwent further radiologic evaluation that was negative for injury, one had a chest tube placed for an occult pneumothorax before exploratory laparotomy, and one patient had a thoracotomy for repair of aortic injury. Chest CT for select patients with an abnormal mediastinal silhouette on CXR would have decreased CT scans by 80% yet still identified patients with an intrathoracic vascular injury. The use of chest CT should be limited to the identification of intrathoracic vascular injuries in the setting of an abnormal mediastinal silhouette on CXR. Therapeutic study, level IV; diagnostic study, level III.

  17. Slot opening optimization of surface mounted permanent magnet motor for cogging torque reduction

    International Nuclear Information System (INIS)

    Abbaszadeh, K.; Rezaee Alam, F.; Teshnehlab, M.

    2012-01-01

    Highlights: ► The slot opening shift method is an efficient method for cogging torque reduction. ► Using slot opening skew method, the trapezoidal waveform of back-emf is maintained. ► Using the conventional slot skewing, the wave shape of back-emf is sinusoidal. ► The novelty of paper is using of air–gap permeance harmonics as objective function. ► Other novelty of this paper is using the different optimization algorithms. - Abstract: In this paper, slot opening skew method is used for cogging torque reduction. A three layer stator model is considered for a six-pole PM-BLDC motor (a PM-BLDC motor with 18-slots, six-poles and length of 5 cm) and then a 2D dual model is extracted for this 3D slot opening skew model. The angular shifts of slot opening position in the first and third layers than middle layer are considered as optimization parameters. Slot opening shape is optimized by using different optimization algorithms, such as, the response surface methodology (RSM), the genetic algorithm (GA) and the particle swarm optimization (PSO). In order to using of GA and PSO, the analytical relationship is derived for the air–gap permeance function. The optimization results of these algorithms are being consistent with each other and are verified with FEA results. The results show the significant reduction of cogging torque about 77%.

  18. Congestion Pricing for Aircraft Pushback Slot Allocation

    Science.gov (United States)

    Zhang, Yaping

    2017-01-01

    In order to optimize aircraft pushback management during rush hour, aircraft pushback slot allocation based on congestion pricing is explored while considering monetary compensation based on the quality of the surface operations. First, the concept of the “external cost of surface congestion” is proposed, and a quantitative study on the external cost is performed. Then, an aircraft pushback slot allocation model for minimizing the total surface cost is established. An improved discrete differential evolution algorithm is also designed. Finally, a simulation is performed on Xinzheng International Airport using the proposed model. By comparing the pushback slot control strategy based on congestion pricing with other strategies, the advantages of the proposed model and algorithm are highlighted. In addition to reducing delays and optimizing the delay distribution, the model and algorithm are better suited for use for actual aircraft pushback management during rush hour. Further, it is also observed they do not result in significant increases in the surface cost. These results confirm the effectiveness and suitability of the proposed model and algorithm. PMID:28114429

  19. Congestion Pricing for Aircraft Pushback Slot Allocation.

    Science.gov (United States)

    Liu, Lihua; Zhang, Yaping; Liu, Lan; Xing, Zhiwei

    2017-01-01

    In order to optimize aircraft pushback management during rush hour, aircraft pushback slot allocation based on congestion pricing is explored while considering monetary compensation based on the quality of the surface operations. First, the concept of the "external cost of surface congestion" is proposed, and a quantitative study on the external cost is performed. Then, an aircraft pushback slot allocation model for minimizing the total surface cost is established. An improved discrete differential evolution algorithm is also designed. Finally, a simulation is performed on Xinzheng International Airport using the proposed model. By comparing the pushback slot control strategy based on congestion pricing with other strategies, the advantages of the proposed model and algorithm are highlighted. In addition to reducing delays and optimizing the delay distribution, the model and algorithm are better suited for use for actual aircraft pushback management during rush hour. Further, it is also observed they do not result in significant increases in the surface cost. These results confirm the effectiveness and suitability of the proposed model and algorithm.

  20. Congestion Pricing for Aircraft Pushback Slot Allocation.

    Directory of Open Access Journals (Sweden)

    Lihua Liu

    Full Text Available In order to optimize aircraft pushback management during rush hour, aircraft pushback slot allocation based on congestion pricing is explored while considering monetary compensation based on the quality of the surface operations. First, the concept of the "external cost of surface congestion" is proposed, and a quantitative study on the external cost is performed. Then, an aircraft pushback slot allocation model for minimizing the total surface cost is established. An improved discrete differential evolution algorithm is also designed. Finally, a simulation is performed on Xinzheng International Airport using the proposed model. By comparing the pushback slot control strategy based on congestion pricing with other strategies, the advantages of the proposed model and algorithm are highlighted. In addition to reducing delays and optimizing the delay distribution, the model and algorithm are better suited for use for actual aircraft pushback management during rush hour. Further, it is also observed they do not result in significant increases in the surface cost. These results confirm the effectiveness and suitability of the proposed model and algorithm.

  1. APPLICATION OF 3D DIGITAL SCANNING AND CAD/CAM SYSTEMS FOR ZIRCONIA INDIRECT RESTORATIONS

    Directory of Open Access Journals (Sweden)

    Mariela Tsanova

    2016-09-01

    Full Text Available Object: To study the CAD/CAM systems’ possibilities for manufacturing Zirconia indirect restorations for restoring vital, vastly devastated teeth. Material and methods: Five patients with vital, heavily destroyed molars were included in the clinical study. The teeth were prepared according to requirements for Zirconia restorations. The impressions were taken with 3D intraoral scanner. The acquired data is transmitted electronically for designing the future restoration by the CAD software. After designing, the restoration is milled by the CAM module and sintered, according to a specially projected programme in accordance with the manufacturing demands. After clinical check-up, the restorations are glazed, air-blasted and cemented. Results: Setting up an individual protocol for working with CAD/CAM systems for restoring vital, severely devastated teeth with Zirconia indirect restorations. A research is made on 3D digital scanning, occlusal and proximal modeling, virtual articulation, analyzing the incline of the cavity walls, the thickness of the restoration and the possibilities for restoring the anatomy of the tooth according to patient’s individual characteristics of the teeth’s arch. Conclusion: The scanning and design process should be evaluated with the same quality control as when using traditional methods. The CAD software requires excellent preparation of the tooth cavity and digital scanning. This is the reason why these systems require a thorough understanding of CAD/CAM calibration and parameters of the product to be clinically acceptable. Once understood, CAD/CAM dentistry can create detailed and accurate restorations that function and perform with great success and longevity.

  2. Slotted cage resonator for high-field magnetic resonance imaging of rodents

    Energy Technology Data Exchange (ETDEWEB)

    Marrufo, O; Vasquez, F; Solis, S E; Rodriguez, A O, E-mail: arog@xanum.uam.mx [Departamento de Ingenieria Electrica, Universidad Autonoma Metropolitana Iztapalapa, Mexico, DF 09340 (Mexico)

    2011-04-20

    A variation of the high-frequency cavity resonator coil was experimentally developed according to the theoretical frame proposed by Mansfield in 1990. Circular slots were used instead of cavities to form the coil endplates and it was called the slotted cage resonator coil. The theoretical principles were validated via a coil equivalent circuit and also experimentally with a coil prototype. The radio frequency magnetic field, B1, produced by several coil configurations was numerically simulated using the finite-element approach to investigate their performances. A transceiver coil, 8 cm long and 7.6 cm in diameter, and composed of 4 circular slots with a 15 mm diameter on both endplates, was built to operate at 300 MHz and quadrature driven. Experimental results obtained with the slotted cage resonator coil were presented and showed very good agreement with the theoretical expectations for the resonant frequency as a function of the coil dimensions and slots. A standard birdcage coil was also built for performance comparison purposes. Phantom images were then acquired to compute the signal-to-noise ratio of both coils showing an important improvement of the slotted cage coil over the birdcage coil. The whole-body images of the mouse were also obtained showing high-quality images. Volume resonator coils can be reliably built following the physical principles of the cavity resonator design for high-field magnetic resonance imaging applications of rodents.

  3. Patient-specific dose estimation for pediatric chest CT

    Energy Technology Data Exchange (ETDEWEB)

    Li Xiang; Samei, Ehsan; Segars, W. Paul; Sturgeon, Gregory M.; Colsher, James G.; Frush, Donald P. [Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 and Department of Radiology, Duke Advanced Imaging Laboratories, Duke University Medical Center, Durham, North Carolina 27705 (United States); Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 (United States); Department of Radiology, Duke Advanced Imaging Laboratories, Duke University Medical Center, Durham, North Carolina 27705 (United States); Department of Physics, Duke University, Durham, North Carolina 27710 (United States); and Department of Biomedical Engineering, Duke University, Durham, North Carolina 27708 (United States); Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 and Department of Radiology, Duke Advanced Imaging Laboratories, Duke University Medical Center, Durham, North Carolina 27705 (United States); Department of Radiology, Duke Advanced Imaging Laboratories, Duke University Medical Center, Durham, North Carolina 27705 (United States); Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 and Global Applied Science Laboratory, GE Healthcare, Waukesha, Wisconsin 53188 (United States); Medical Physics Graduate Program, Duke University, Durham, North Carolina 27705 and Department of Radiology, Division of Pediatric Radiology, Duke University Medical Center, Durham North Carolina 27710 (United States)

    2008-12-15

    Current methods for organ and effective dose estimations in pediatric CT are largely patient generic. Physical phantoms and computer models have only been developed for standard/limited patient sizes at discrete ages (e.g., 0, 1, 5, 10, 15 years old) and do not reflect the variability of patient anatomy and body habitus within the same size/age group. In this investigation, full-body computer models of seven pediatric patients in the same size/protocol group (weight: 11.9-18.2 kg) were created based on the patients' actual multi-detector array CT (MDCT) data. Organs and structures in the scan coverage were individually segmented. Other organs and structures were created by morphing existing adult models (developed from visible human data) to match the framework defined by the segmented organs, referencing the organ volume and anthropometry data in ICRP Publication 89. Organ and effective dose of these patients from a chest MDCT scan protocol (64 slice LightSpeed VCT scanner, 120 kVp, 70 or 75 mA, 0.4 s gantry rotation period, pitch of 1.375, 20 mm beam collimation, and small body scan field-of-view) was calculated using a Monte Carlo program previously developed and validated to simulate radiation transport in the same CT system. The seven patients had normalized effective dose of 3.7-5.3 mSv/100 mAs (coefficient of variation: 10.8%). Normalized lung dose and heart dose were 10.4-12.6 mGy/100 mAs and 11.2-13.3 mGy/100 mAs, respectively. Organ dose variations across the patients were generally small for large organs in the scan coverage (<7%), but large for small organs in the scan coverage (9%-18%) and for partially or indirectly exposed organs (11%-77%). Normalized effective dose correlated weakly with body weight (correlation coefficient: r=-0.80). Normalized lung dose and heart dose correlated strongly with mid-chest equivalent diameter (lung: r=-0.99, heart: r=-0.93); these strong correlation relationships can be used to estimate patient-specific organ

  4. Scanning laser optical tomography for in toto imaging of the murine cochlea.

    Directory of Open Access Journals (Sweden)

    Lena Nolte

    Full Text Available The mammalian cochlea is a complex macroscopic structure due to its helical shape and the microscopic arrangements of the individual layers of cells. To improve the outcomes of hearing restoration in deaf patients, it is important to understand the anatomic structure and composition of the cochlea ex vivo. Hitherto, only one histological technique based on confocal laser scanning microscopy and optical clearing has been developed for in toto optical imaging of the murine cochlea. However, with a growing size of the specimen, e.g., human cochlea, this technique reaches its limitations. Here, we demonstrate scanning laser optical tomography (SLOT as a valuable imaging technique to visualize the murine cochlea in toto without any physical slicing. This technique can also be applied in larger specimens up to cm3 such as the human cochlea. Furthermore, immunolabeling allows visualization of inner hair cells (otoferlin or spiral ganglion cells (neurofilament within the whole cochlea. After image reconstruction, the 3D dataset was used for digital segmentation of the labeled region. As a result, quantitative analysis of position, length and curvature of the labeled region was possible. This is of high interest in order to understand the interaction of cochlear implants (CI and cells in more detail.

  5. A 51-Year-Old Woman With an Increasing Chest Wall Mass Years After Resection of an Early Stage Lung Cancer.

    Science.gov (United States)

    Dhakal, Ajay; Chen, Hongbin; Dexter, Elisabeth U

    2017-12-01

    A 51-year-old woman was found to have a new 14 × 6 mm soft tissue mass under the right serratus muscle on a CT scan of the chest performed for routine surveillance due to her history of stage I lung cancer. A follow-up CT scan performed 4 months later showed that the mass had increased in size to 22 × 8 mm. The patient presents to the oncology clinic to discuss the results of the CT scan. She has no pain or swelling on the right lateral chest and no cough, fever, or shortness of breath. She is at her baseline health with good appetite and functional status. Copyright © 2017 American College of Chest Physicians. Published by Elsevier Inc. All rights reserved.

  6. Application of low-dose radiation protocols in survey CT scans

    International Nuclear Information System (INIS)

    Fu Qiang; Liu Ting; Lu Tao; Xu Ke; Zhang Lin

    2009-01-01

    Objective: To characterize the protocols with low-dose radiation in survey CT scans for localization. Methods: Eighty standard adult patients, head and body phantoms were recruited. Default protocols provided by operator's manual setting were that all the tube voltage for head, chest, abdomen and lumbar was 120 kV; the tube currents were 20,10,20 and 40 mA, respectively. Values of kV and mA in the low-dose experiments were optimized according to the device options. For chest and abdomen, the tube position were compared between default (0 degree) and 180 degree. Phantoms were scanned with above protocols, and the radiation doses were measured respectively. Paired t-test were used for comparisons of standard deviation in CT value, noise and exposure surface dose (ESD) between group with default protocols and group with optimized protocols. Results: The optimized protocols in low-dose CT survey scans were 80 kV, 10 mA for head, 80 kV, 10 mA for chest, 80 kV, 10 mA for abdomen and 100 kV, 10 mA for lumbar. The values of ESD for phantom scan in default and optimized protocols were 0.38 mGy/0.16 mGy in head, 0.30 mGy/0.20 mGy in chest, 0.74 mGy/0.30 mGy in abdomen and 0.81 mGy/0.44 mGy in lumbar, respectively. Compared with default protocols, the optimized protocols reduced the radiation doses 59%, 33%, 59% and 46% in head, chest, abdomen and lumbar. When tube position changed from 0 degree to 180 degree, the ESD were 0.24 mGy/0.20 mGy for chest; 0.37 mGy/0.30 mGy for abdomen, and the radiation doses were reduced 20% and 17%. Conclusion: A certain amount of image noise is increased in low-dose protocols, but image quality is still acceptable without problem in CT localization. The reduction of radiation dose and the radiation harm to patients are the superiority. (authors)

  7. Evaluation of two styles of slotted, flat-head screws

    International Nuclear Information System (INIS)

    Reeves, C.A. Jr.; Johnson, W.B.

    1979-01-01

    A series of torque tests were performed to evaluate the relative merits of two different flat-head screws fabricated from a uranium--6% niobium alloy. The screws tested were machined with both normal, straight-through slots in the head and with slots having radiused bottoms. Test results indicate that both designs easily surpass the required 20-inch-pound-proof torque

  8. Chest X-Ray

    Medline Plus

    Full Text Available ... about chest radiography also known as chest x-rays. Chest x-rays are the most commonly performed x-ray exams and use a very small dose of ... of the inside of the chest. A chest x-ray is used to evaluate the lungs, heart and ...

  9. Development of high power models of four-slot Annular Coupled Structure

    International Nuclear Information System (INIS)

    Kageyama, T.; Morozumi, Y.; Yoshino, K.; Yamazaki, Y.

    1994-01-01

    A π/2-mode standing-wave linac (f=1.296 GHz) of an Annular Coupled Structure (ACS) has been developed for the 1-GeV proton linac of the Japanese Hadron Project (JHP). This ACS has four coupling slots between accelerating and coupling cells in order to suppress higher order mode mixing with the π/2 coupling mode. High-β(β=v/c=0.78) and low-β(0.52) prototypes were constructed and tested up to each design RF power. Concerning the effect of the coupling slots on the fields of a coupled-cavity linac, it was found that the slot configuration of the side-coupled structure (SCS) tilts the accelerating field. On the other hand, the four-slot configuration of the ACS gives an almost axially symmetric accelerating field to the beam. (author)

  10. AIRFIX: the first digital postoperative chest tube airflowmetry--a novel method to quantify air leakage after lung resection.

    Science.gov (United States)

    Anegg, Udo; Lindenmann, Jorg; Matzi, Veronika; Mujkic, Dzenana; Maier, Alfred; Fritz, Lukas; Smolle-Jüttner, Freyja Maria

    2006-06-01

    Prolonged air leak after pulmonary resection is a common complication and a major limiting factor for early discharge from hospital. Currently there is little consensus on its management. The aim of this study was to develop and evaluate a measuring device which allows a simple digital bed-side quantification of air-leaks compatible to standard thoracic drainage systems. The measuring device (AIRFIX) is based upon a 'mass airflow' sensor with a specially designed software package that is connected to a thoracic suction drainage system. Its efficacy in detecting pulmonary air-leaks was evaluated in a series of 204 patients; all postoperative measurements were done under standardized conditions; the patients were asked to cough, to take a deep breath, to breathe out against the resistance of a flutter valve, to keep breath and to breathe normally. As standard parameters, the leakage per breath or cough (ml/b) as well as the leakage per minute (ml/min) were displayed and recorded on the computer. Air-leaks within a range of 0.25-45 ml/b and 5-900 ml/min were found. Removal of the chest tubes was done when leakage volume on Heimlich valve was less than 1.0 ml/b or 20 ml/min. After drain removal based upon the data from chest tube airflowmetry none of the patients needed re-drainage due to pneumothorax. The AIRFIX device for bed-side quantification of air-leaks has proved to be very simple and helpful in diagnosis and management of air-leaks after lung surgery, permitting drain removal without tentative clamping.

  11. High slot utilization systems for electric machines

    Science.gov (United States)

    Hsu, John S

    2009-06-23

    Two new High Slot Utilization (HSU) Systems for electric machines enable the use of form wound coils that have the highest fill factor and the best use of magnetic materials. The epoxy/resin/curing treatment ensures the mechanical strength of the assembly of teeth, core, and coils. In addition, the first HSU system allows the coil layers to be moved inside the slots for the assembly purpose. The second system uses the slided-in teeth instead of the plugged-in teeth. The power density of the electric machine that uses either system can reach its highest limit.

  12. Validity, reliability, and reproducibility of linear measurements on digital models obtained from intraoral and cone-beam computed tomography scans of alginate impressions

    NARCIS (Netherlands)

    Wiranto, Matthew G.; Engelbrecht, W. Petrie; Nolthenius, Heleen E. Tutein; van der Meer, W. Joerd; Ren, Yijin

    INTRODUCTION: Digital 3-dimensional models are widely used for orthodontic diagnosis. The aim of this study was to assess the validity, reliability, and reproducibility of digital models obtained from the Lava Chairside Oral scanner (3M ESPE, Seefeld, Germany) and cone-beam computed tomography scans

  13. Radionuclide scanning

    International Nuclear Information System (INIS)

    Shapiro, B.

    1986-01-01

    Radionuclide scanning is the production of images of normal and diseased tissues and organs by means of the gamma-ray emissions from radiopharmaceutical agents having specific distributions in the body. The gamma rays are detected at the body surface by a variety of instruments that convert the invisible rays into visible patterns representing the distribution of the radionuclide in the body. The patterns, or images, obtained can be interpreted to provide or to aid diagnoses, to follow the course of disease, and to monitor the management of various illnesses. Scanning is a sensitive technique, but its specificity may be low when interpreted alone. To be used most successfully, radionuclide scanning must be interpreted in conjunction with other techniques, such as bone radiographs with bone scans, chest radiographs with lung scans, and ultrasonic studies with thyroid scans. Interpretation is also enhanced by providing pertinent clinical information because the distribution of radiopharmaceutical agents can be altered by drugs and by various procedures besides physiologic and pathologic conditions. Discussion of the patient with the radionuclide scanning specialist prior to the study and review of the results with that specialist after the study are beneficial

  14. Comparison of gated radionuclide scans and chest radiographs. Assessment of left ventricular impairment in patients with coronary disease.

    Science.gov (United States)

    Bianco, J A; Reinke, D B; Makey, D G; Shafer, R B

    1980-03-01

    Diagnostic efficacy of gated cardiac blood pool imaging was studied in 41 consecutive patients with LV ejection fractions (LVEF) less than or equal to 0.50. Eighty percent of patients were receiving therapy for LV failure at the time of the study. All patients had documented coronary-artery disease (CAD). Chest x-ray films were interpreted blindly by a senior radiologist. Cardiothoracic ratio of less than or equal to 0.50 was recorded as normal. Radionuclide assessment of LV function contributes importantly to the diagnostic and screening value of chest x-ray films. Patients with coronary disease and clinical evidence of heart failure should have radioisotopic studies even if chest x-ray film findings are normal. In patients with coronary artery disease and enlarged LV on chest films, radionuclide study of left ventricular performance aids in defining LV impairment, and in the prognostication of subsequent clinical course.

  15. Digital X-Radiography Scanning Laboratory

    Data.gov (United States)

    Federal Laboratory Consortium — FUNCTION: Generates digital X-radiographic images of sediment cores that portray density variations, sediment stratigraphy, bioturbation, and inclusions.DESCRIPTION:...

  16. Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging.

    Science.gov (United States)

    Tanaka, Rie

    2016-07-01

    Dynamic chest radiography is a flat-panel detector (FPD)-based functional X-ray imaging, which is performed as an additional examination in chest radiography. The large field of view (FOV) of FPDs permits real-time observation of the entire lungs and simultaneous right-and-left evaluation of diaphragm kinetics. Most importantly, dynamic chest radiography provides pulmonary ventilation and circulation findings as slight changes in pixel value even without the use of contrast media; the interpretation is challenging and crucial for a better understanding of pulmonary function. The basic concept was proposed in the 1980s; however, it was not realized until the 2010s because of technical limitations. Dynamic FPDs and advanced digital image processing played a key role for clinical application of dynamic chest radiography. Pulmonary ventilation and circulation can be quantified and visualized for the diagnosis of pulmonary diseases. Dynamic chest radiography can be deployed as a simple and rapid means of functional imaging in both routine and emergency medicine. Here, we focus on the evaluation of pulmonary ventilation and circulation. This review article describes the basic mechanism of imaging findings according to pulmonary/circulation physiology, followed by imaging procedures, analysis method, and diagnostic performance of dynamic chest radiography.

  17. Improvement The Transmission Efficiency For Wireless Packet Communication Systems Using Automatic Control for power And Time Slot Width Of Slotted Non persistent ISMA Protocol

    Directory of Open Access Journals (Sweden)

    Saad M. Hardan

    2013-05-01

    Full Text Available In packed communication systems which use a protocol, the protocol should perform the allocation of channels such that the transmission channel is used efficiently. Efficiency is usually measured in terms of channel throughput and the average transmission  delay. The Slotted Nonpersistent ISMA protocol is one of random access protocols used in packed communication systems. In this research a Slotted Nonpersistent ISMA protocol Model with automatic control for power and time slot is proposed. the suggested algorithm enable the base station(access point to control  the protocol time slot length and  transmission power in a dynamic way to control the normalized propagation delay d and to maintain all the uplink signals in the limit of captured power threshold (capture ratio in order to control the  throughput and the average transmission delay of the communication system by an automatic method. the computer simulation results  confirm the activity of the  proposed algorithm for increasing the  throughput and decreasing the average transmission delay by an accepted ratios.

  18. The digital Emily project: achieving a photorealistic digital actor.

    Science.gov (United States)

    Alexander, Oleg; Rogers, Mike; Lambeth, William; Chiang, Jen-Yuan; Ma, Wan-Chun; Wang, Chuan-Chang; Debevec, Paul

    2010-01-01

    The Digital Emily Project uses advanced face scanning, character rigging, performance capture, and compositing to achieve one of the world's first photorealistic digital facial performances. The project scanned the geometry and reflectance of actress Emily O'Brien's face in 33 poses, showing different emotions, gaze directions, and lip formations in a light stage. These high-resolution scans-accurate to skin pores and fine wrinkles-became the basis for building a blendshape-based facial-animation rig whose expressions closely matched the scans. The blendshape rig drove displacement maps to add dynamic surface detail. A video-based facial animation system animated the face according to the performance in a reference video, and the digital face was tracked onto the video's motion and rendered under the same illumination. The result was a realistic 3D digital facial performance credited as one of the first to cross the "uncanny valley" between animated and fully human performances.

  19. Bead beavior in a non-continues slot die coating regime

    NARCIS (Netherlands)

    Langen, A.; Senes, A.; Vries, I. de; Groen, P.

    2015-01-01

    Sheet-to-sheet slot die coating is a crucial intermediate technique to bring organic and large area electronics (e.g. OLED and Flexible OPV) towards mass roll-to-roll productions. The coating on substrate has to be uniform within nanometers. Additional settings are added to the slot die coater for a

  20. Fluid Collection in the Right Lateral Portion of the Superior Aortic Recess Mimicking a Right Mediastinal Mass: Assessment with Chest Posterior Anterior and MDCT

    International Nuclear Information System (INIS)

    Shn, Dong Rock; Ryu, Dae Shick; Park, Man Soo; Jung, Seung Mun; Ahn, Jae Hong; Lee, Jong Hyeog; Choi, Soo Jung

    2012-01-01

    We observed patients in whom the fluid collection in the right lateral portion of the superior aortic recess on computed tomography (CT) scans mimicked a right anterior mediastinal mass on chest PA radiographs. The purpose of this study was to assess chest PA and CT features of these patients. All chest PA radiographs and CT scans in 9 patients were reviewed by two radiologists on a consensus basis; for the presence of pleural effusion, pulmonary edema and heart size on chest PA radiographs. For the portion of the fluid collection in the superior aortic recess (SAR), a connection between the right lateral portion of the SAR (rSAR) and posterior portion of the SAR (pSAR) on CT scans, and the distance between the right lateral margin of the rSAR and the right lateral margin of the superior vena cava. Fluid collection in the rSAR on CT scans caused a right anterior mediastinal mass or a bulging contour on chest PA radiographs in all women patients. All patients showed cardiomegaly, five patients had pleural effusion, and two patients had mild pulmonary edema. Further, eight patients showed a connection between the rSAR and the pSAR. The characteristic features of these patients are the right anterior mediastinal mass-like opacity due to fluid collection in the rSAR, are bulging contour with a smooth margin and cardiomegaly regardless of pulmonary edema on the chest PA radiographs, and fluid connection between the rSAR and the pSAR on CT scans

  1. Fluid Collection in the Right Lateral Portion of the Superior Aortic Recess Mimicking a Right Mediastinal Mass: Assessment with Chest Posterior Anterior and MDCT

    Energy Technology Data Exchange (ETDEWEB)

    Shn, Dong Rock; Ryu, Dae Shick; Park, Man Soo; Jung, Seung Mun; Ahn, Jae Hong; Lee, Jong Hyeog; Choi, Soo Jung [Dept. of Radiology, Gangneung Asan Hospital, College of Medicine, University of Ulsan, Gangneung (Korea, Republic of)

    2012-09-15

    We observed patients in whom the fluid collection in the right lateral portion of the superior aortic recess on computed tomography (CT) scans mimicked a right anterior mediastinal mass on chest PA radiographs. The purpose of this study was to assess chest PA and CT features of these patients. All chest PA radiographs and CT scans in 9 patients were reviewed by two radiologists on a consensus basis; for the presence of pleural effusion, pulmonary edema and heart size on chest PA radiographs. For the portion of the fluid collection in the superior aortic recess (SAR), a connection between the right lateral portion of the SAR (rSAR) and posterior portion of the SAR (pSAR) on CT scans, and the distance between the right lateral margin of the rSAR and the right lateral margin of the superior vena cava. Fluid collection in the rSAR on CT scans caused a right anterior mediastinal mass or a bulging contour on chest PA radiographs in all women patients. All patients showed cardiomegaly, five patients had pleural effusion, and two patients had mild pulmonary edema. Further, eight patients showed a connection between the rSAR and the pSAR. The characteristic features of these patients are the right anterior mediastinal mass-like opacity due to fluid collection in the rSAR, are bulging contour with a smooth margin and cardiomegaly regardless of pulmonary edema on the chest PA radiographs, and fluid connection between the rSAR and the pSAR on CT scans.

  2. Development of a pediatric chest phantom for dosimetry in computerized tomography

    International Nuclear Information System (INIS)

    Aburjaile, W.N.; Mourão, A.P.; Oliveira, F.A.

    2017-01-01

    Computed Tomography (CT) has promoted a significant increase in the dose absorbed by patients due to the diagnosis. Therefore, it is indispensable to improve protocols, using smaller doses, without impairing the diagnostic quality of the image. The risks of stochastic effects are greater for children due to tissue radiosensitivity coupled with longer life expectancy. In the work, a cylindrical simulator was used, representing an adult thorax, made of polymethylmethacrylate, and a second simulator object of the same material was developed in oblong format including the axillary regions based on the dimensions of the pediatric patient's 8-year-old chest. A comparative study was performed between chest scans performed on two CT equipment in different radiodiagnostic services. The central slice of the two simulating objects was irradiated successively and, using a pencil-type ionization chamber, the absorbed dose was measured at five different points of each simulating object. From the measurements, weighted Dose and Volumetric Dose Index (CK, PMMA, vol) values were obtained for the 10 cm sweep of the central region of the object, in helical mode. The scans were performed using the chest acquisition protocols used by radiodiagnostic services, both for a supply voltage of the 120 kV X-ray tube. The study allowed to compare the variation of absorbed dose between patients with distinct chest volumes and the patient dose variation between two devices when used for the generation of images with the same diagnostic objective

  3. Temporal digital subtraction radiography with a personal computer digital workstation

    International Nuclear Information System (INIS)

    Kircos, L.; Holt, W.; Khademi, J.

    1990-01-01

    Technique have been developed and implemented on a personal computer (PC)-based digital workstation to accomplish temporal digital subtraction radiography (TDSR). TDSR is useful in recording radiologic change over time. Thus, this technique is useful not only for monitoring chronic disease processes but also for monitoring the temporal course of interventional therapies. A PC-based digital workstation was developed on a PC386 platform with add-in hardware and software. Image acquisition, storage, and processing was accomplished using 512 x 512 x 8- or 12-bit frame grabber. Software and hardware were developed to accomplish image orientation, registration, gray scale compensation, subtraction, and enhancement. Temporal radiographs of the jaws were made in a fixed and reproducible orientation between the x-ray source and image receptor enabling TDSR. Temporal changes secondary to chronic periodontal disease, osseointegration of endosseous implants, and wound healing were demonstrated. Use of TDSR for chest imaging was also demonstrated with identification of small, subtle focal masses that were not apparent with routine viewing. The large amount of radiologic information in images of the jaws and chest may obfuscate subtle changes that TDSR seems to identify. TDSR appears to be useful as a tool to record temporal and subtle changes in radiologic images

  4. Development of an economical and ecological optimized multi slot fish bypass

    International Nuclear Information System (INIS)

    Tauber, M.; Mader, H.

    2009-01-01

    The European Union's Water Framework Directive (WFD), which came into force in 2000, represents a new framework governing water policies. The ecological objectives of the directive are sometimes in conflict with the historical use of waterbodies for hydropower generation. This paper discussed the mitigation of negative economical effects caused by outfitting hydropower plants with fish migration facilities claimed by the WFD. In particular, it reviewed the new modified version of the Vertical Slot Fishpass which reduces water discharge while maintaining or even enhancing the biological migration of aquatic species. Reducing the flow rate in the fish bypass results in more available water for hydropower generation, which is economically advantageous for operators. The available water resource is simply used more efficiently. The bypass was modified by adding additional slot baffles and guide walls in order to induce intended losses at roughness obstacles and to obtain isolated roughness current in the multi slot area. A Froude similarity model was used to obtain the results of the first phase of this project. Different arrangements of the slot baffles and locations were measured. The results of all trials revealed that the newly developed multi slot variants have clear advantages over conventional vertical slot fish passages. Discharge was reduced by up to 34 per cent while still maintaining the same water levels in the pools. As such, the general flow velocities were also reduced by about 10 per cent, thereby achieving the primary objective of the project. 7 refs., 18 figs

  5. A model of gas flow with friction in a slotted seal

    Directory of Open Access Journals (Sweden)

    Joachimiak Damian

    2016-09-01

    Full Text Available The paper discusses thermodynamic phenomena accompanying the flow of gas in a slotted seal. The analysis of the gas flow has been described based on an irreversible adiabatic transformation. A model based on the equation of total enthalpy balance has been proposed. The iterative process of the model aims at obtaining such a gas temperature distribution that will fulfill the continuity equation. The model allows for dissipation of the kinetic energy into friction heat by making use of the Blasius equation to determine the friction coefficient. Within the works, experimental research has been performed of the gas flow in a slotted seal of slot height 2 mm. Based on the experimental data, the equation of local friction coefficient was modified with a correction parameter. This parameter was described with the function of pressure ratio to obtain a mass flow of the value from the experiment. The reason for taking up of this problem is the absence of high accuracy models for calculating the gas flow in slotted seals. The proposed model allows an accurate determination of the mass flow in a slotted seal based on the geometry and gas initial and final parameters.

  6. Digital radiography

    International Nuclear Information System (INIS)

    Brody, W.R.

    1984-01-01

    Digital Radiography begins with an orderly introduction to the fundamental concepts of digital imaging. The entire X-ray digital imagining system is described, from an overall characterization of image quality to specific components required for a digital radiographic system. Because subtraction is central to digital radiographic systems, the author details the use of various subtraction methods for image enhancement. Complex concepts are illustrated with numerous examples and presented in terms that can readily be understood by physicians without an advanced mathematics background. The second part of the book discusses implementations and applications of digital imagining systems based on area and scanned detector technologies. This section includes thorough coverage of digital fluoroscopy, scanned projection radiography, and film-based digital imaging systems, and features a state-of-the-art synopsis of the applications of digital subtraction angiography. The book concludes with a timely assessment of anticipated technological advances

  7. Reconstruction of Chest Wall by Cryopreserved Sternal Allograft after Resection of Aneurysmal Bone Cyst of Sternum

    Directory of Open Access Journals (Sweden)

    Kambiz Sheikhy

    2017-01-01

    Full Text Available A 20-year-old female was referred to our hospital due to deformity and bulging in anterior aspect of chest wall in sternal area. Chest X-ray and CT scan confirmed a large mass with destruction of sternum. Pathologic diagnosis after incisional biopsy was compatible with aneurysmal bone cyst. We resected sternum completely and reconstructed large anterior defect by a cryopreserved sternal allograft. In follow-up of patient there was no unstability of chest wall with good cosmetic result.

  8. Rotational scanning and multiple-spot focusing through a multimode fiber based on digital optical phase conjugation

    Science.gov (United States)

    Ma, Chaojie; Di, Jianglei; Li, Ying; Xiao, Fajun; Zhang, Jiwei; Liu, Kaihui; Bai, Xuedong; Zhao, Jianlin

    2018-06-01

    We demonstrate, for the first time, the rotational memory effect of a multimode fiber (MMF) based on digital optical phase conjugation (DOPC) to achieve multiple-spot focusing. An implementation interferometer is used to address the challenging alignments in DOPC. By rotating the acquired phase conjugate pattern, rotational scanning through a MMF could be achieved by recording a single off-axis hologram. The generation of two focal spots through a MMF is also demonstrated by combining the rotational memory effect with the superposition principle. The results may be useful for ultrafast scanning imaging and optical manipulation of multiple objects through a MMF.

  9. Chest X-Ray

    Medline Plus

    Full Text Available ... talk with you about chest radiography also known as chest x-rays. Chest x-rays are the ... treatment for a variety of lung conditions such as pneumonia, emphysema and cancer. A chest x-ray ...

  10. The Effects of In-Office Reconditioning on the Slot Dimensions and Static Frictional Resistance of Stainless Steel Brackets.

    Science.gov (United States)

    Iluru, Rohini; Nellore, Chaitanya; Karnati, Praveen Kumar Reddy; Thalapaneni, Ashok Kumar; Myla, Vijay Bhaskar; Ramyasree, Konda; Prasad, Mandava

    2016-01-01

    Orthodontists are commonly faced with the decision of what to do with loose brackets, and with inaccurately located brackets that need repositioning during treatment. One solution is to recycle the brackets. The potential effects of reconditioning a bracket are dependent upon many factors which may result in physical changes like alteration in slot tolerance, which may influence sliding mechanics by affecting frictional resistance. To study and compare the dimensional changes in the bracket slot width and depth in reconditioned brackets from unused brackets under scanning electronic microscope and to study and compare any consequent effects on the static frictional resistance of stainless steel brackets after reconditioning and in unused brackets. Dentarum manufactured 90 stainless steel central incisors edgewise brackets of size 0.22 X 0.030″ inch and 0° tip and 0°angulation were taken. 60 samples for measuring frictional resistance and 30 samples for measuring slot dimensions. Ortho organizers manufactured stainless steel arch wires 0.019 X 0.025″ straight lengths 60 in number were considered for measuring static frictional resistance. The mean slot width and depth of new brackets were 0.0251″ and 0.0471″, which exceeded the manufacturers reported nominal size of 0.022″ X 0.030″, by 0.003″ and 0.017″. The reconditioned brackets demonstrated a further increase in mean slot width and depth to 0.028″ and 0.0518″ that is by 0.0035″ and 0.0047″ which is statistically significant (p=0.001, 0.002). The mean static frictional forces of the reconditioned brackets was nearly similar to that of new brackets that is 0.3167N for reconditioned brackets and 0.2613 N for new brackets. Although the reconditioning process results in physical changes to bracket structure this does not appear to result in significant effect on ex-vivo static frictional resistance.

  11. Chest X-Ray

    Medline Plus

    Full Text Available ... by Image/Video Gallery Your Radiologist Explains Chest X-ray Transcript Welcome to Radiology Info dot org! Hello, ... d like to talk with you about chest radiography also known as chest x-rays. Chest x- ...

  12. FLAIL CHEST

    Directory of Open Access Journals (Sweden)

    Anton Crnjac

    2003-12-01

    Full Text Available Background. Major thoracic trauma is consistent with high mortality rate because of associated injuries of vital thoracic organs and dangerous complications. The flail chest occurs after disruption of the skeletal continuity of chest wall and demands because of its pathophysiological complexity rapid and accurate diagnosis and treatment.Conclusions. Basic pathophysiological mechanism of the flail chest is respiratory distress, which is provoked by pulmonary contusions and paradoxical chest wall motion. The treatment should be pointed to improvement and support of respiratory functions and include aggressive pain control, pulmonary physiotherapy and selective mechanical ventilation. Views about operative fixation of the flail chest are still controversial. Neither mortality rate neither long-term disability are improved after operative fixation.

  13. Digital Dentistry — Digital Impression and CAD/CAM System Applications

    Directory of Open Access Journals (Sweden)

    Gabor Alin-Gabriel

    2017-03-01

    Full Text Available Digital imprint and computer-aided design/computer-aided manufacture (CAD/CAM systems offer several benefits compared to traditional techniques. The use of a CAD/CAM system to scan preparations and generate restorations in-office, removes a second appointment for the patient. The existence of precision benefits in using complete systems and chairside scanning systems, has been proven. CAD/CAM restorations have a good longevity and meet the accepted clinical parameters. New digital impression methods are presently accessible, and before long, the long-awaited goal of sparing patients of one the most unpleasant practices in clinical dentistry, acquiring dental impressions, will be exchanged by intraoral digital scanning. CAD/CAM systems existing nowadays, can feed data through accurate digital scans created from plaster models, straight to manufacturing systems that can shape ceramic or resin restorations with no requirement of a physical copy of the prepared, adjacent, and antagonist teeth.

  14. Modeling fiber motion in a pulp pressure screen: the effect of slot shape

    International Nuclear Information System (INIS)

    Dong, S.; Salcudean, M.; Gartshore, I.

    2003-01-01

    A pressure screen is a piece of equipment in the pulp and paper industry used either to remove contaminants from the pulp suspension or to separate fibers having different properties. Contaminants such as fiber bundles, bark and plastic specks are introduced when fibers are separated from the wood by mechanical or chemical pulping processes. Contaminants significantly affect the strength and smoothness of the paper and must be removed before the final paper is produced. The screen plate is a critical part of the pressure screen and its design is the key to screen performance. This paper uses a new and comprehensive CFD simulation tool to examine the flow and fiber behavior in a single slot screen having any reasonable slot shape. -This simulation tool includes three coupled models: first, the flow model solves the Reynolds Averaged Navier-Stokes (RANS) equation using the standard k - ε turbulence model to predict the flow field in the equipment. Second, a three-dimensional flexible fiber model is used to track the fiber trajectory in the screen. Third, a very general wall model is used to deal with the case when a fiber touches the equipment wall. The simulated results show that the slot shape has a critical influence on fiber behavior and screen performance. Three general slot shapes were investigated: the smooth slot, the step-step contour slot and slope-slope contour slot. Of these the slope-slope contour slot provides the best passage for the fibers with a length of 1mm and 3mm. (author)

  15. Chest X-Ray

    Medline Plus

    Full Text Available ... by Image/Video Gallery Your Radiologist Explains Chest X-ray Transcript Welcome to Radiology Info dot org! Hello, ... you about chest radiography also known as chest x-rays. Chest x-rays are the most commonly performed ...

  16. Automated detection of analyzable metaphase chromosome cells depicted on scanned digital microscopic images

    Science.gov (United States)

    Qiu, Yuchen; Wang, Xingwei; Chen, Xiaodong; Li, Yuhua; Liu, Hong; Li, Shibo; Zheng, Bin

    2010-02-01

    Visually searching for analyzable metaphase chromosome cells under microscopes is quite time-consuming and difficult. To improve detection efficiency, consistency, and diagnostic accuracy, an automated microscopic image scanning system was developed and tested to directly acquire digital images with sufficient spatial resolution for clinical diagnosis. A computer-aided detection (CAD) scheme was also developed and integrated into the image scanning system to search for and detect the regions of interest (ROI) that contain analyzable metaphase chromosome cells in the large volume of scanned images acquired from one specimen. Thus, the cytogeneticists only need to observe and interpret the limited number of ROIs. In this study, the high-resolution microscopic image scanning and CAD performance was investigated and evaluated using nine sets of images scanned from either bone marrow (three) or blood (six) specimens for diagnosis of leukemia. The automated CAD-selection results were compared with the visual selection. In the experiment, the cytogeneticists first visually searched for the analyzable metaphase chromosome cells from specimens under microscopes. The specimens were also automated scanned and followed by applying the CAD scheme to detect and save ROIs containing analyzable cells while deleting the others. The automated selected ROIs were then examined by a panel of three cytogeneticists. From the scanned images, CAD selected more analyzable cells than initially visual examinations of the cytogeneticists in both blood and bone marrow specimens. In general, CAD had higher performance in analyzing blood specimens. Even in three bone marrow specimens, CAD selected 50, 22, 9 ROIs, respectively. Except matching with the initially visual selection of 9, 7, and 5 analyzable cells in these three specimens, the cytogeneticists also selected 41, 15 and 4 new analyzable cells, which were missed in initially visual searching. This experiment showed the feasibility of

  17. Substernal thyroid carcinoma detected by 67Ga scan in a patient with normal 131I scan

    International Nuclear Information System (INIS)

    Kim, e.E.; Maruyama, Y.; Deland, F.H.

    1978-01-01

    A patient with a superior mediastinal mass on an admission chest radiograph was initially evaluated by an 131 I thyroid scan which failed to demonstrate a substernal thyroid. However, the tomographic 67 Ga scan clearly showed an abnormal uptake in the area corresponding to the mass lesion on radiographic examination. Subsequent resection and biopsy of the substernal mass revealed a poorly differentiated follicular carcinoma with foci of anaplastic carcinoma. The differential diagnosis of the anterior mediastinal mass and the usefullness of the tomographic gallium scan are briefly discussed

  18. Study on wake structure characteristics of a slotted micro-ramp with large-eddy simulation

    Energy Technology Data Exchange (ETDEWEB)

    Dong, Xiangrui; Chen, Yaohui; Dong, Gang; Liu, Yixin, E-mail: cyh873@163.com [National Key Laboratory of Transient Physics, Nanjing University of Science and Technology, Nanjing, 210094 (China)

    2017-06-15

    In this paper, a novel slotted ramp-type micro vortex generator (slotted micro-ramp) for flow separation control is simulated in the supersonic flow of Ma = 1.5, based on large eddy simulation combined with the finite volume method. The wake structure characteristics and control mechanisms of both slotted and standard micro-ramps are presented and discussed. The results show that the wake of standard micro-ramp includes a primary counter-rotating streamwise vortex pair, a train of vortex rings, and secondary vortices. The slotted micro-ramp has more complicated wake structures, which contain a confluent counter-rotating streamwise vortex pair and additional streamwise vortices, with the same rotation generated by slot and the vortex rings enveloping the vortex pair. The additional vortices generated by the slot of the micro-ramp can mix with the primary counter-rotating vortex pair, extend the life time, and strengthen the vortex intensity of primary vortex pair. Moreover, the slot can effectively alleviate, or even eliminate the backflow and decrease the profile drag induced by the standard micro-ramp, therefore improving the efficiency of separation control. (paper)

  19. Radiological imaging of the neonatal chest. 2. rev. ed.

    Energy Technology Data Exchange (ETDEWEB)

    Donoghue, Veronica (ed.) [Children' s University Hospital, Dublin (Ireland). Dept. of Radiology; National Maternity Hospital, Dublin (Ireland)

    2008-07-01

    This second, revised edition of Radiological Imaging of the Neonatal Chest provides a comprehensive and up-to-date discussion of the subject. It is written primarily from the point of view of the paediatric radiologist but will be of particular interest to all antenatal ultrasonographers, neonatologists, paediatric cardiologists, paediatricians and paediatric surgeons. It includes an update on clinical management and appraises the advantages of the various techniques available to image the newborn chest. There is particular emphasis on the impact of recent therapeutic advances on imaging findings. Extensive consideration is given to both antenatal and postnatal imaging of congenital chest malformations, as well as to controversies regarding the postnatal management of asymptomatic infants with these anomalies. There are dedicated chapters on upper airway problems, infection and congenital heart disease, with special emphasis on the current role of magnetic resonance imaging, computed tomography and interventional therapy. There is also a chapter devoted to computed radiography and digital radiography. This book contains important information for all those involved in caring for the neonate. (orig.)

  20. Slot technique - an alternative method of scatter reduction in radiography

    International Nuclear Information System (INIS)

    Panzer, W.; Widenmann, L.

    1983-01-01

    The most common method of scatter reduction in radiography is the use of an antiscatter grid. Its disadvantage is the absorption of a certain percentage of primary radiation in the lead strips of the grid and the fact that due to the limited thickness of the lead strips their scatter absorption is also limited. A possibility for avoiding this disadvantage is offered by the so-called slot technique, ie, the successive exposure of the subject with a narrow fan beam provided by slots in rather thick lead plates. The results of a comparison between grid and slot technique regarding dose to the patient, scatter reduction, image quality and the effect of automatic exposure control are reported. (author)

  1. Cusp-Gun Sixth-Harmonic Slotted Gyrotron

    Science.gov (United States)

    Stutzman, R. C.; McDermott, D. B.; Hirata Luhmann, Y., Jr.; Gallagher, D. A.; Spencer, T. A.

    2000-10-01

    A high-harmonic slotted gyrotron has been constructed at UC Davis to be driven by a 70 kV, 3.5 A, axis-encircling electron beam from a Northrop Grumman Cusp gun. The 94 GHz, slotted sixth-harmonic gyrotron is predicted to generate 50 kW with an efficiency of 20%. Using the profile of the adiabatic field reversal from the UC Davis superconducting test-magnet, EGUN simulations predict that an axis-encircling electron beam will be generated with an axial velocity spread of Δ v_z/v_z=10% for the desired velocity ratio of α =v_z/v_z=1.5. The design will also be presented for an 8th-harmonic W-band gyrotron whose magnetic field can be supplied by a lightweight permanent magnet.

  2. Tri-Band CPW-Fed Stub-Loaded Slot Antenna Design for WLAN/WiMAX Applications

    Science.gov (United States)

    Li, Jianxing; Guo, Jianying; He, Bin; Zhang, Anxue; Liu, Qing Huo

    2016-11-01

    A novel uniplanar CPW-fed tri-band stub-loaded slot antenna is proposed for wireless local area network (WLAN) and worldwide interoperability for microwave access (WiMAX) applications. Dual resonant modes were effectively excited in the upper band by using two identical pairs of slot stubs and parasitic slots symmetrically along the arms of a traditional CPW-fed slot dipole, achieving a much wider bandwidth. The middle band was realized by the fundamental mode of the slot dipole. To obtain the lower band, two identical inverted-L-shaped open-ended slots were symmetrically etched in the ground plane. A prototype was fabricated and measured, showing that tri-band operation with 10-dB return loss bandwidths of 150 MHz from 2.375 to 2.525 GHz, 725 MHz from 3.075 to 3.8 GHz, and 1.9 GHz from 5.0 to 6.9 GHz has been achieved. Details of the antenna design as well as the measured and simulated results are presented and discussed.

  3. Radiology illustrated. Chest radiology

    International Nuclear Information System (INIS)

    Lee, Kyung Soo; Han, Joungho; Chung, Man Pyo; Jeong, Yeon Joo

    2014-01-01

    Pattern approach to the diagnosis of lung diseases based on CT scan appearances. Guide to quick and reliable differential diagnosis. CT-pathology correlation. Emphasis on state-of-the-art MDCT. The purpose of this atlas is to illustrate how to achieve reliable diagnoses when confronted by the different abnormalities, or ''disease patterns'', that may be visualized on CT scans of the chest. The task of pattern recognition has been greatly facilitated by the advent of multidetector CT (MDCT), and the focus of the book is very much on the role of state-of-the-art MDCT. A wide range of disease patterns and distributions are covered, with emphasis on the typical imaging characteristics of the various focal and diffuse lung diseases. In addition, clinical information relevant to differential diagnosis is provided and the underlying gross and microscopic pathology is depicted, permitting CT-pathology correlation. The entire information relevant to each disease pattern is also tabulated for ease of reference. This book will be an invaluable handy tool that will enable the reader to quickly and easily reach a diagnosis appropriate to the pattern of lung abnormality identified on CT scans.

  4. Image quality in conventional chest radiography. Evaluation using the postprocessing tool Diamond View

    International Nuclear Information System (INIS)

    Niemann, Tilo; Reisinger, Clemens; Rau, Philipp; Schwarz, Jochen; Ruis-Lopez, Laura; Bongartz, Georg

    2010-01-01

    The objective of this work was to evaluate the influence of the postprocessing tool Diamond View (Siemens AG Medical Solutions, Germany) on image quality in conventional chest radiography. Evaluation of image quality remains a challenge in conventional radiography. Based on the European Commission quality criteria we evaluated the improvement of image quality when applying the new postprocessing tool Diamond View (Siemens AG Medical solutions, Germany) to conventional chest radiographs. Three different readers prospectively evaluated 102 digital image pairs of chest radiographs. Statistical analysis was performed with a p value <0.05 considered as significant. Images were evaluated on basis of the modified imaging Quality Criteria by the Commission of the European Communities. Each of the 11 image quality criteria was evaluated separately using a five point classification. Statistical analysis showed an overall tendency for improved image quality for Diamond View (DV) for all criteria. Significant differences could be found in most of the criteria. In conclusion DV improves image quality in conventional chest radiographs.

  5. Lung involvement quantification in chest radiographs; Quantificacao de comprometimento pulmonar em radiografias de torax

    Energy Technology Data Exchange (ETDEWEB)

    Giacomini, Guilherme; Alvarez, Matheus; Oliveira, Marcela de; Miranda, Jose Ricardo A. [Universidade Estadual Paulista Julio de Mesquita Filho (UNESP), Botucatu, SP (Brazil). Instituto de Biociencias. Departamento de Fisica e Biofisica; Pina, Diana R.; Pereira, Paulo C.M.; Ribeiro, Sergio M., E-mail: giacomini@ibb.unesp.br [Universidade Estadual Paulista Julio de Mesquita Filho (UNESP), Botucatu, SP (Brazil). Faculdade de Medicina. Departamento de Doencas Tropicais e Diagnostico por Imagem

    2014-12-15

    Tuberculosis (TB) caused by Mycobacterium tuberculosis, is an infectious disease which remains a global health problem. The chest radiography is the commonly method employed to assess the TB's evolution. The methods for quantification of abnormalities of chest are usually performed on CT scans (CT). This quantification is important to assess the TB evolution and treatment and comparing different treatments. However, precise quantification is not feasible for the amount of CT scans required. The purpose of this work is to develop a methodology for quantification of lung damage caused by TB through chest radiographs. It was developed an algorithm for computational processing of exams in Matlab, which creates a lungs' 3D representation, with compromised dilated regions inside. The quantification of lung lesions was also made for the same patients through CT scans. The measurements from the two methods were compared and resulting in strong correlation. Applying statistical Bland and Altman, all samples were within the limits of agreement, with a confidence interval of 95%. The results showed an average variation of around 13% between the two quantification methods. The results suggest the effectiveness and applicability of the method developed, providing better risk-benefit to the patient and cost-benefit ratio for the institution. (author)

  6. Chest X-Ray

    Medline Plus

    Full Text Available ... I’d like to talk with you about chest radiography also known as chest x-rays. Chest x-rays are the most ... far outweighs any risk. For more information about chest x-rays, visit Radiology Info dot org. Thank you for your time! ...

  7. Comparison of digital selenium radiography with an analog screen-film system in the diagnostic process of pneumoconiosis according to ILO classification

    International Nuclear Information System (INIS)

    Zaehringer, M.; Winnekendonk, G.; Gossmann, A.; Krueger, K.; Krug, B.

    2001-01-01

    Purpose: The aim of the study was to determine the diagnostic value of digital selenium radiography in patients with pneumoconiosis. For this purpose chest X-rays by digital selenium radiography and analog screen-film system were compared according to the ILO classification of pneumoconiosis. Method: After approval of the study by the local ethic commission and the Federal German Office for Radiation Protection 50 patients were subjected to X-rays by digital selenium radiography (Thoravision; Philips Medical Systems, Hamburg, Germany) and analog screen-film system of the same day within the scope of an industrial medicine preventive checkup. Four investigators rated the chest X-rays according to the ILO classification of pneumoconiosis. Results: The findings demonstrated by chest X-rays according to ILO classification were rated similar by digital selenium radiography and analog screen film systems. Image quality of the digital pictures was rated significantly better. Conclusion: The use of digital selenium radiography in evaluating chest X-rays according to the ILO classification does not result in over- or underestimation of pulmonary pathologies. Hence, in the diagnosis of pneumoconiosis, digital selenium radiography can replace the tested analog screen-film system. (orig.) [de

  8. EFFECTS OF SLOTTED BLADING ON SECONDARY FLOW IN HIGHLY LOADED COMPRESSOR CASCADE

    Directory of Open Access Journals (Sweden)

    RAMZI MDOUKI

    2013-10-01

    Full Text Available With the aim to increase allowable blade loadings and enlarge stable operating range in highly loaded compressor, this work is carried out in order to explore the potential of passive control via slotted bladings in linear cascade configurations under both design and stall conditions. Through an extensive 2D-numerical study, the effects of location, width and slope of slots were analysed and the best configuration was identified. Based on the optimal slot, the 3D aerodynamic performances of cascade were studied and the influence of slotted blading to control endwall flow was investigated. Both 2D and 3D calculations are performed on steady RANS solver with standard k-epsilon turbulence model and low Mach number regime. The total loss coefficient, turning angle and flow visualizations on the blade and end-wall surfaces are adopted to describe the different configurations. The obtained results show, for 2D situation, that a maximum of 28.3% reduction in loss coefficient had been reached and the flow turning was increased with approximately 5°. Concerning 3D flow fields the slots marked their benefit at large incoming flow angles which delays the separation on both end wall and blade suction surface at mid span. However, at design conditions, the slotted blades are not able to control secondary flows near the wall and so, lose their potential.

  9. FLOW VISUALIZATION OF RECTANGULAR SLOT AIR JET IMPINGEMENT ON FLAT SURFACES

    OpenAIRE

    Satheesha V *1, B. K. Muralidhra2, Abhilash N3, C. K. Umesh4

    2018-01-01

    Jet impingement near the mid-chord of the gas turbine blade is treated as a flat plate. Experimental and numerical investigations are carried out for a single slot air jet impinging on flat surface for two different rectangular slots of dimension (3mm x 65 mm) and (5mm x 65 mm). Experimentation is done to study the flow pattern topography on the flat target plate, with varying the flow rate from 20 LPM to 50 LPM by varying the nozzle to plate distance from 9 mm to 24 mm for slot jet of 3mm an...

  10. Chest imaging with dual-energy substraction digital tomosynthesis

    International Nuclear Information System (INIS)

    Sone, S.; Kasuga, T.; Sakai, F.; Hirano, H.; Kubo, K.; Morimoto, M.; Takemura, K.; Hosoba, M.

    1993-01-01

    Dual-energy subtraction digital tomosynthesis with pulsed X-ray and rapid kV switching was used to examine calcifications in pulmonary lesions. The digital tomosynthesis system used included a conventional fluororadiographic TV unit with linear tomographic capabilities, a high resolution videocamera, and an image processing unit. Low-voltage, high voltage, and soft tissue subtracted or bone subtracted tomograms of any desired layer height were reconstructed from the image data acquired during a single tomographic swing. Calcifications, as well as their characteristics and distribution in pulmonary lesions, were clearly shown. The images also permitted discrimination of calcifications from dense fibrotic lesions. This technique was effective in demonstrating calcifications together with a solitary mass or disseminated nodules. (orig.)

  11. Average chest wall thickness at two anatomic locations in trauma patients.

    Science.gov (United States)

    Schroeder, Elizabeth; Valdez, Carrie; Krauthamer, Andres; Khati, Nadia; Rasmus, Jessica; Amdur, Richard; Brindle, Kathleen; Sarani, Babak

    2013-09-01

    Needle thoracostomy is the emergent treatment for tension pneumothorax. This procedure is commonly done using a 4.5cm catheter, and the optimal site for chest wall puncture is controversial. We hypothesize that needle thoracostomy cannot be performed using this catheter length irrespective of the site chosen in either gender. A retrospective review of all chest computed tomography (CT) scans obtained on trauma patients from January 1, 2011 to December 31, 2011 was performed. Patients aged 18 and 80 years were included and patients whose chest wall thickness exceeded the boundary of the images acquired were excluded. Chest wall thickness was measured at the 2nd intercostal (ICS), midclavicular line (MCL) and the 5th ICS, anterior axillary line (AAL). Injury severity score (ISS), chest wall thickness, and body mass index (BMI) were analyzed. 201 patients were included, 54% male. Average (SD) BMI was 26 (7)kg/m(2). The average chest wall thickness in the overall cohort was 4.08 (1.4)cm at the 2nd ICS/MCL and 4.55 (1.7)cm at the 5th ICS/AAL. 29% of the overall cohort (27 male and 32 female) had a chest wall thickness greater than 4.5cm at the 2nd ICS/MCL and 45% (54 male and 36 female) had a chest wall thickness greater than 4.5cm at the 5th ICS/AAL. There was no significant interaction between gender and chest wall thickness at either site. BMI was positively associated with chest wall thickness at both the 2nd and 5th ICS/AAL. A 4.5cm catheter is inadequate for needle thoracostomy in most patients regardless of puncture site or gender. Copyright © 2013 Elsevier Ltd. All rights reserved.

  12. Increased inspiratory pressure for reduction of atelectasis in children anesthetized for CT scan

    International Nuclear Information System (INIS)

    Sargent, Michael A.; Jamieson, Douglas H.; McEachern, Anita M.; Blackstock, Derek

    2002-01-01

    Background: Atelectasis is more frequent and more severe in children anesthetized for CT scan than it is in children sedated for CT scan.Objective: To determine the effect of increased inspiratory pressure on atelectasis during chest CT in anesthetized children. Materials and methods: Atelectasis on chest CT was assessed by two observers in three groups of patients. Group A comprised 13 children (26 lungs) anesthetized at inspiratory pressures up to and including 25 cm H 2 O. Group B included 11 children anesthetized at inspiratory pressures ≥30 cm H 2 O. Group C included 8 children under deep sedation. Results: Atelectasis was significantly more severe in group A than in groups B and C. There was no significant difference between groups B and C. Conclusion: An inspiratory pressure of 30 cm H 2 O is recommended for children anesthetized for CT scan of the chest. (orig.)

  13. Low-dosage helical CT applications for chest medical checkup and lung cancer screening

    International Nuclear Information System (INIS)

    Wang Ping; Cui Fa; Liang Huanqing; Zheng Minfei

    2005-01-01

    Objective: A discussion on low-dosage helical CT applications on chest medical checkup and lung cancer screening. Methods: On the 100 chest medical check up with three different of protocols, including standard-dosage (the tube current was 230 mAs) were compared with low-dose (tube current was 50 mAs or 30 mAs). Results: Low-dosage helical CT scan provides excellent images. In 100 chest medical checkup, 39 nodules or masses were revealed, enlarged lymph node was noted in 1 case; emphysema or bullae was demonstrated in 3 segments; thickening of bronchial wall was shown in 2 cases; and localized pleural thickening was found in 1 case. Conclusion: In chest checkup or lung cancer screening low-dosage helical CT (tube current 30 mAs) will not only guarantee image quality but also reduce the radiation dose during the examination. (authors)

  14. Diagnostic Accuracy of Ultrasonography and Radiography in Initial Evaluation of Chest Trauma Patients.

    Science.gov (United States)

    Vafaei, Ali; Hatamabadi, Hamid Reza; Heidary, Kamran; Alimohammadi, Hosein; Tarbiyat, Mohammad

    2016-01-01

    Application of chest radiography for all multiple trauma patients is associated with a significant increase in total costs, exposure to radiation, and overcrowding of the emergency department. Ultrasound has been introduced as an alternative diagnostic tool in this regard. The aim of the present study is to compare the diagnostic accuracy of chest ultrasonography and radiography in detection of traumatic intrathoracic injuries. In the present prospective cross-sectional study, patients with traumatic intrathoracic injuries, who were referred to the emergency department from December 2013 to December 2014, were assessed. The patients underwent bedside ultrasound, radiographic and computed tomography (CT) scan examinations based on ATLS recommendations. Screening performance characteristics of ultrasonography and radiography were compared using SPSS 21.0. Chest CT scan was considered as gold standard. 152 chest trauma patients with a mean age of 31.4 ± 13.8 years (range: 4 ‒ 67), were enrolled (77.6% male). Chest CT scan showed pulmonary contusion in 48 (31.6%) patients, hemothorax in 29 (19.1%), and pneumothorax in 55 (36.2%) cases. Area under the ROC curve of ultrasonography in detection of pneumothorax, hemothorax, and pulmonary contusion were 0.91 (95% CI: 0.86‒0.96), 0.86 (95% CI: 0.78‒0.94), and 0.80 (95% CI: 0.736‒0.88), respectively. Area under the ROC curve of radiography was 0.80 (95% CI: 0.736‒0.87) for detection of pneumothorax, 0.77 (95% CI: 0.68‒0.86) for hemothorax, and 0.58 (95% CI: 0.5‒0.67) for pulmonary contusion. Comparison of areas under the ROC curve declared the significant superiority of ultrasonography in detection of pneumothorax (p = 0.02) and pulmonary contusion (p < 0.001). However, the diagnostic value of the two tests was equal in detection of hemothorax (p = 0.08). The results of the present study showed that ultrasonography is preferable to radiography in the initial evaluation of patients with traumatic injuries to the

  15. Diagnostic Accuracy of Ultrasonography and Radiography in Initial Evaluation of Chest Trauma Patients

    Directory of Open Access Journals (Sweden)

    Ali Vafaei

    2016-01-01

    Full Text Available Introduction: Application of chest radiography for all multiple trauma patients is associated with a significant increase in total costs, exposure to radiation, and overcrowding of the emergency department. Ultrasound has been introduced as an alternative diagnostic tool in this regard. The aim of the present study is to compare the diagnostic accuracy of chest ultrasonography and radiography in detection of traumatic intrathoracic injuries. Methods: In the present prospective cross-sectional study, patients with traumatic intrathoracic injuries, who were referred to the emergency department from December 2013 to December 2014, were assessed. The patients underwent bedside ultrasound, radiographic and computed tomography (CT scan examinations based on ATLS recommendations. Screening performance characteristics of ultrasonography and radiography were compared using SPSS 21.0. Chest CT scan was considered as gold standard. Results: 152 chest trauma patients with a mean age of 31.4 ± 13.8 years (range: 4 ‒ 67, were enrolled (77.6% male. Chest CT scan showed pulmonary contusion in 48 (31.6% patients, hemothorax in 29 (19.1%, and pneumothorax in 55 (36.2% cases. Area under the ROC curve of ultrasonography in detection of pneumothorax, hemothorax, and pulmonary contusion were 0.91 (95% CI: 0.86‒0.96, 0.86 (95% CI: 0.78‒0.94, and 0.80 (95% CI: 0.736‒0.88, respectively. Area under the ROC curve of radiography was 0.80 (95% CI: 0.736‒0.87 for detection of pneumothorax, 0.77 (95% CI: 0.68‒0.86 for hemothorax, and 0.58 (95% CI: 0.5‒0.67 for pulmonary contusion. Comparison of areas under the ROC curve declared the significant superiority of ultrasonography in detection of pneumothorax (p = 0.02 and pulmonary contusion (p < 0.001. However, the diagnostic value of the two tests was equal in detection of hemothorax (p = 0.08. Conclusion: The results of the present study showed that ultrasonography is preferable to radiography in the initial

  16. Increased opacity of left pericardiac area on chest radiograph : correlation with CT findings and its frequency

    International Nuclear Information System (INIS)

    Lee, Seung Ik; Kim, Jeung Sook; Kwak, Jin Young; Ryu, Chang Woo; Yoon, Sam Hyun

    1998-01-01

    The purpose of this study was to evaluate the incidence of extrapericardial fat in the left cardiac border, and with regard to left extrapericardial fat, to correlate chest radiographs with CT scans. This study involved 132 consecutive patients who underwent chest PA and lateral radiographs, and chest CT scans. According to the results of chest PA radiograph, patients were divided into three groups: those with a clear left cardiac border; those with an indistinct left cardiac border; and those with an indistinct left cardiac border with increased density; cardiophrenic angle, as seen on lateral radiograph, the presence of increased density in the anterior cardiophrenic angle, as seen on lateral radiographs was evaluated. On the basis of the results of CT scanning, patients were classified into four groups according to the amount of left extrapericardial fat : negative, minimum, moderate, and maximum. Left extrapericardial fat, as seen on CT, was correlated with the conspicuity of left cardiac border seen on PA radiograph and the presence of increased density in the anterior cardiophrenic angle, as seen on lateral radiograph. The conspicuousness of the left cardiac border, as seen on PA chest radiograph, correlated with the presence of left extrapericardial fat, as seen on CT, and was related to the amount of left extrapericardial fat. Increased density of the anterior cardiophrenic angle, as seen on lateral radiographs, correlated with the presence of left extrapericardial fat on CT, but the absence of increased density on lateral radiograph corresponds to the absence or a minimal amount of left extrapericardial fat, as seen on CT. (author). 8 refs., 4 tabs., 3 figs

  17. Film-screen digital radiography

    International Nuclear Information System (INIS)

    Schwenker, R.P.; Eger, H.

    1985-01-01

    The excellent performance of the digital film-screen system as a receptor for projection radiographic data is discussed. An experimental system for obtaining high quality digital radiographic data by laser scanning radiographic films is described. This system is being used to evaluate the clinical utility of various digital image processing algorithms. Future plans include an investigation of quantitative analysis of projection radiographic data. Digital data obtained by film scanning can be used with digital image archiving and communications systems. (author)

  18. Intraoral scan bodies in implant dentistry: A systematic review.

    Science.gov (United States)

    Mizumoto, Ryan M; Yilmaz, Burak

    2018-04-05

    Intraoral scan body (ISB) design is highly variable and its role in the digital workflow and accuracy of digital impressions is not well understood. The purpose of this systematic review was to determine the relevant reports pertaining to ISBs with regard to design and accuracy and to describe their evolution and role in the digital dentistry workflow. Special attention was placed on their key features in relation to intraoral scanning technology and the digitization process. A MEDLINE/PubMed search was performed to identify relevant reports pertaining to ISB usage in dentistry. This search included but was not limited to scan body features and design, scan body accuracy, and scan body techniques and the role of ISBs in computer-aided design and computer-aided manufacturing (CAD-CAM) processes. Commercially available scan bodies were examined, and a patient situation was shown highlighting the use of ISBs in the digital workflow. Deficiencies in the reports were found regarding various scan body topics, including ISB features/design, accuracy, and the role of ISBs in CAD-CAM processes. ISBs are complex implant-positioning-transfer devices that play an essential role in the digital workflow and fabrication of accurately fitting implant-supported restorations. With scanner technology rapidly evolving and becoming more widespread, future studies are needed and should be directed toward all parts of the digital workflow when using ISBs. By understanding the basic components of ISBs and how they relate to digital scanning and CAD-CAM technology, more emphasis may be placed on their importance and usage in the digital workflow to ensure accurate transfer of implant position to the virtual and analog definitive cast. Efforts should be made by clinicians to identify an optimal ISB design in relation to the specific intraoral scanning technology being used. Copyright © 2017 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights

  19. Small angle slot divertor concept for long pulse advanced tokamaks

    Science.gov (United States)

    Guo, H. Y.; Sang, C. F.; Stangeby, P. C.; Lao, L. L.; Taylor, T. S.; Thomas, D. M.

    2017-04-01

    SOLPS-EIRENE edge code analysis shows that a gas-tight slot divertor geometry with a small-angle (glancing-incidence) target, named the small angle slot (SAS) divertor, can achieve cold, dissipative/detached divertor conditions at relatively low values of plasma density at the outside midplane separatrix. SAS exhibits the following key features: (1) strong enhancement of the buildup of neutral density in a localized region near the plasma strike point on the divertor target; (2) spreading of the cooling front across the divertor target with the slot gradually flaring out from the strike point, thus effectively reducing both heat flux and erosion on the entire divertor target surface. Such a divertor may potentially provide a power and particle handling solution for long pulse advanced tokamaks.

  20. FEM Analysis of Brushless DC Servomotor with Fractional Number of Slots per Pole

    Directory of Open Access Journals (Sweden)

    BALUTA, G.

    2014-02-01

    Full Text Available The authors present in this paper the analysis with Finite Element Method (FEM of the magnetic circuit for a Brushless DC servomotor with fractional number of slots/pole (9 slots and 10 poles. For this purpose, FEMM 4.2 software package was used for the analysis. To obtain the waveforms of Back-ElectroMotive Forces (BEMFs, electromagnetic and cogging torque for servomotor a program in LUA scripting language (integrated into interactive shell of FEMM4.2 has been created. A comparation with a structure with integer number of slots/pole (18 slots and 6 poles was also realized. The analysis results prove that the structure chosen is an optimal solution: sinusoidal waveforms of BEMFs, improved electromagnetic torque and reduced cogging torque. Therefore, the operating characteristics of the servomotor with 9/10 slots/poles manufactured by Sistem Euroteh Company and included in an integrated electrical drives system are presented in this paper.

  1. Cogging Torque Reduction by Slot-Opening Shift for Permanent Magnet Machines

    DEFF Research Database (Denmark)

    Liu, Ting; Huang, Shoudao; Gao, Jian

    2013-01-01

    In this paper, an effective cogging torque reduction method based on shifting the slot-openings is presented. Stator teeth are divided into groups and proper slot-opening shift is applied for each group. The cogging torque can then be greatly reduced while the back-EMF waveforms remain symmetrical...

  2. A simple method of digitizing analog scintigrams for quantification and digital archiving

    International Nuclear Information System (INIS)

    Schramm, M.; Kaempfer, B.; Wolf, H.; Clausen, M.; Wendhausen, H.; Henze, E.

    1993-01-01

    This study was undertaken to evaluate a quick, reliable and cheap method of digitizing analog scintigrams. 40 whole-body bone scintigrams were obtained simultaneously in analog and genuine digital format. The analog scans on X-ray film were then digitized seecondarily by three different methods: 300 dpi flatbed scanning, high-resolution camera scanning and camcorder recording. A simple exposure approach using a light box, a cheap camcorder, a PC and image grabber hard- and software proved to be optimal. Visual interpretation showed no differences in clinical findings when comparing the analog images with their secondarily digitized counterparts. To test the possibility of quantification, 126 equivalent ROIs were drawn both in the genuine digital and the secondarily digitized images. Comparing the ROI count to whole-body count percentage of the corresponding ROIs showed the correlation to be linear. The evaluation of phantom studies showed the linear correlation to be true within a wide activity range. Thus, secondary digitalization of analog scintigrams is an easy, cheap and reliable method of archiving images and allows secondary digital quantification. (orig.) [de

  3. [A simple method of digitizing analog scintigrams for quantification and digital archiving].

    Science.gov (United States)

    Schramm, M; Kämpfer, B; Wolf, H; Clausen, M; Wendhausen, H; Henze, E

    1993-02-01

    This study was undertaken to evaluate a quick, reliable and cheap method of digitizing analog scintigrams. 40 whole-body bone scintigrams were obtained simultaneously in analog and genuine digital format. The analog scans on x-ray film were then digitized secondarily by three different methods: 300 dpi flat-bed scanning, high-resolution camera scanning and camcorder recording. A simple exposure approach using a light box, a cheap camcorder, a PC and image grabber hard- and software proved to be optimal. Visual interpretation showed no differences in clinical findings when comparing the analog images with their secondarily digitized counterparts. To test the possibility of quantification, 126 equivalent ROIs were drawn both in the genuine digital and the secondarily digitized images. Comparing the ROI count to whole-body count percentage of the corresponding ROIs showed the correlation to be linear. The evaluation of phantom studies showed the linear correlation to be true within a wide activity range. Thus, secondary digitalization of analog scintigrams is an easy, cheap and reliable method of archiving images and allows secondary digital quantification.

  4. Pulmonary complications of crack cocaine use: high-resolution computed tomography of the chest

    International Nuclear Information System (INIS)

    Mancano, Alexandre

    2008-01-01

    Here, we report high-resolution computed tomography (HRCT) findings in a patient who developed sudden hemoptysis, dyspnea and chest pain after smoking crack cocaine. Chest X-rays showed consolidations, primarily in the upper lobes, and HRCT scans showed ground glass attenuation opacities, consolidations and air-space nodules. A follow-up CT, after drug use discontinuation and administration of corticosteroids, showed partial resolution of pulmonary lesions and the appearance of cavitations. Clinical, imaging and laboratory findings led to a diagnosis of 'crack lung'. (author)

  5. Hydraulic performance of a low specific speed centrifugal pump with Spanwise-Slotted Blades

    International Nuclear Information System (INIS)

    Ye, D X; Li, H; Wang, Y

    2013-01-01

    The hydraulic efficiency of a low specific speed centrifugal pump is low because of the long and narrow meridian flow passage, and the severe disk friction. Spanwise slotted blade flow control technology has been applied to the low specific speed centrifugal pump. This paper concluded that spanwise slotted blades can improve the pump performance in both experiments and simulations. In order to study the influence to the impeller and volute by spanwise slotted blade, impeller efficiency and volute efficiency were defined. The minimum volute efficiency and the maximum pump efficiency appear at the same time in the design flow condition in the unsteady simulation. The mechanism of spanwise slotted blade flow control technology should be researched furthermore

  6. Effects of tube potential and scatter rejection on image quality and effective dose in digital chest X-ray examination: An anthropomorphic phantom study

    Energy Technology Data Exchange (ETDEWEB)

    Shaw, D.J., E-mail: daniel.shaw@christie.nhs.uk [Diagnostic Radiology, Department of Medical Physics and Engineering, Leeds Teaching Hospitals, Leeds General Infirmary, Great George Street, Leeds LS1 3EX (United Kingdom); Crawshaw, I. [Diagnostic X-ray Department, York Teaching Hospital NHS Foundation Trust, The York Hospital, Wigginton Road, York YO31 8HE (United Kingdom); Rimmer, S. D. [Diagnostic Radiology, Department of Medical Physics and Engineering, Leeds Teaching Hospitals, Leeds General Infirmary, Great George Street, Leeds LS1 3EX (United Kingdom)

    2013-11-15

    Objectives: The purpose of this study was to investigate the effects of tube potential and scatter rejection techniques on image quality of digital posteroanterior (PA) chest radiographs. Methods: An anthropomorphic phantom was imaged using a range of tube potentials (81–125 kV{sub p}) without scatter rejection, with an anti-scatter grid, and using a 10 cm air gap. Images were anonymised and randomised before being evaluated using a visual graded analysis (VGA) method. Results: The effects of tube potential on image quality were found to be negligible (p > 0.63) for the flat panel detector (FPD). Decreased image quality (p = 0.031) was noted for 125 kV{sub p} relative to 109 kV{sub p}, though no difference was noted for any of the other potentials (p > 0.398) for computed radiography (CR). Both scatter rejection techniques improved image quality (p < 0.01). For FPD imaging the anti-scatter grid offered slightly improved image quality relative to the air gap (p = 0.038) but this was not seen for CR (p = 0.404). Conclusions: For FPD chest imaging of the anthropomorphic phantom there was no dependence of image quality on tube potential. Scatter rejection improved image quality, with the anti-scatter grid giving greater improvements than an air-gap, but at the expense of increased effective dose. CR imaging of the chest phantom demonstrated negligible dependence on tube potential except at 125 kV{sub p}. Scatter rejection improved image quality, but with no difference found between techniques. The air-gap resulted in a smaller increase in effective dose than the anti-scatter grid and would be the preferred scatter rejection technique.

  7. Improving Abnormality Detection on Chest Radiography Using Game-Like Reinforcement Mechanics.

    Science.gov (United States)

    Chen, Po-Hao; Roth, Howard; Galperin-Aizenberg, Maya; Ruutiainen, Alexander T; Gefter, Warren; Cook, Tessa S

    2017-11-01

    Despite their increasing prevalence, online textbooks, question banks, and digital references focus primarily on explicit knowledge. Implicit skills such as abnormality detection require repeated practice on clinical service and have few digital substitutes. Using mechanics traditionally deployed in video games such as clearly defined goals, rapid-fire levels, and narrow time constraints may be an effective way to teach implicit skills. We created a freely available, online module to evaluate the ability of individuals to differentiate between normal and abnormal chest radiographs by implementing mechanics, including instantaneous feedback, rapid-fire cases, and 15-second timers. Volunteer subjects completed the modules and were separated based on formal experience with chest radiography. Performance between training and testing sets were measured for each group, and a survey was administered after each session. The module contained 74 cases and took approximately 20 minutes to complete. Thirty-two cases were normal radiographs and 56 cases were abnormal. Of the 60 volunteers recruited, 25 were "never trained" and 35 were "previously trained." "Never trained" users scored 21.9 out of 37 during training and 24.0 out of 37 during testing (59.1% vs 64.9%, P value online module may improve the abnormality detection rates of novice interpreters of chest radiography, although experienced interpreters are less likely to derive similar benefits. Users reviewed the educational module favorably. Copyright © 2017 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

  8. Digital hilar tomography. Comparison with conventional technique

    International Nuclear Information System (INIS)

    Schaefer, C.B.; Braunschweig, R.; Teufl, F.; Kaiser, W.; Claussen, C.D.

    1993-01-01

    The aim of the following study was to compare conventional hilar tomography and digital hilar tomography. 20 patients were examined both with conventional and digital hilar tomography using the same tomographic technique and the identical exposure dose. All patients underwent computed tomography of the chest as a golden standard. The digital technique, especially the edge-enhanced image version, showed superior image quality. ROC-analysis by 4 readers found equal diagnostic performance without any statistical difference. Digital hilar tomography shows a superior and constant image quality and lowers the rate of re-exposure. Therefore, digital hilar tomography is the preferable method. (orig.) [de

  9. Realistic Free-Spins Features Increase Preference for Slot Machines.

    Science.gov (United States)

    Taylor, Lorance F; Macaskill, Anne C; Hunt, Maree J

    2017-06-01

    Despite increasing research into how the structural characteristics of slot machines influence gambling behaviour there have been no experimental investigations into the effect of free-spins bonus features-a structural characteristic that is commonly central to the design of slot machines. This series of three experiments investigated the free-spins feature using slot machine simulations to determine whether participants allocate more wagers to a machine with free spins, and, which components of free-spins features drive this preference. In each experiment, participants were exposed to two computer-simulated slot machines-one with a free-spins feature or similar bonus feature and one without. Participants then completed a testing phase where they could freely switch between the two machines. In Experiment 1, participants did not prefer the machine with a simple free-spins feature. In Experiment 2 the free-spins feature incorporated additional elements such as sounds, animations, and an increased win frequency; participants preferred to gamble on this machine. The Experiment 3 "bonus feature" machine resembled the free spins machine in Experiment 2 except spins were not free; participants showed a clear preference for this machine also. These findings indicate that (1) free-spins features have a major influence over machine choice and (2) the "freeness" of the free-spins bonus features is not an important driver of preference, contrary to self-report and interview research with gamblers.

  10. 14 CFR 93.226 - Allocation of slots in low-demand periods.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 2 2010-01-01 2010-01-01 false Allocation of slots in low-demand periods. 93.226 Section 93.226 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF... low-demand periods. (a) If there are available slots in the following time periods and there are no...

  11. Elevated CPW-Fed Slotted Microstrip Antenna for Ultra-Wideband Application

    Directory of Open Access Journals (Sweden)

    Chandan Kumar Ghosh

    2012-01-01

    Full Text Available Elevated-coplanar-waveguide- (ECPW- fed microstrip antenna with inverted “G” slots in the back conductor is presented. It is modeled and analyzed for the application of multiple frequency bands. The changes in radiation and the transmission characteristics are investigated by the introduction of the slots in two different positions at the ground plane (back conductor. The proposed antenna without slots exhibits a stop band from 2.55 GHz to 4.25 GHz while introducing two slots on the back conductor, two adjacent poles appear at central frequencies of 3.0 GHz and 3.9 GHz, respectively, and the antenna shows the ultra-wideband (UWB characteristics. The first pole appears at the central frequency of 3.0 GHz and covers the band width of 950 MHz, and the second pole exists at a central frequency of 3.90 GHz covering a bandwidth of 750 MHz. Experimental result shows that impedance bandwidth of 129% (S11<-10 dB is well achieved when the antenna is excited with both slots. Compared to most of the previously reported ECPW structures, the impedance bandwidth of this antenna is increased and also the size of the antenna becomes smaller and more suitable for many wireless applications like PCS (1850–1990 MHz, WLAN (2.4–2.484 GHz, WiMAX (2.5–2.69 GHz and 5.15–5.85 GHz, and also X-band communication.

  12. CT dose profiles and MSAD calculation in a chest phantom

    International Nuclear Information System (INIS)

    Oliveira, Bruno Beraldo; Silva, Teogenes Augusto da

    2011-01-01

    For optimizing patient doses in computed tomography (CT), the Brazilian legislation has only established diagnostic reference levels (DRLs) in terms of Multiple Scan Average Dose (MSAD) in a typical adult as a quality control parameter for CT scanners. Compliance with the DRLs can be verified by measuring the Computed Tomography Air Kerma Index with a calibrated pencil ionization chamber or by obtaining the dose distribution in CT scans. An analysis of the quality of five CT scanners in Belo Horizonte was done in terms of dose profile of chest scans and MSAD determinations. Measurements were done with rod shape lithium fluoride thermoluminescent dosimeters (TLD-100) distributed in cylinders positioned in peripheral and central regions of a polymethylmethacrylate chest phantom. The peripheral regions presented higher dose values. The longitudinal dose variation can be observed and the maximum dose was recorded at the edges of the phantom at the midpoint of the longitudinal axis. The MSAD results were in according to the DRL of 25 mGy established by Brazilian legislation. The results contribute to disseminate to hospitals and radiologists the proper procedure to use the thermoluminescent dosimeters for the calculation of the MSAD from the CT dose profiles and to notice the compliance with the DRLs. (author)

  13. Numerical study on non-locally reacting behavior of nacelle liners incorporating drainage slots

    Science.gov (United States)

    Chen, Chao; Li, Xiaodong; Thiele, Frank

    2018-06-01

    For acoustic liners used in current commercial nacelles, in order to prevent any liquid accumulating in the resonators, drainage slots are incorporated on the partition walls between closely packed cavities. Recently, an experimental study conducted by Busse-Gerstengarbe et al. shown that the cell interaction introduced by drainage slots causes an additional dissipation peak which increases with the size of the slot. However, the variation of damping process due to drainage slots is still not fully understood. Therefore, a numerical study based on computational aeroacoustic methods is carried out to investigate the mechanism of the changed attenuation characteristics due to drainage slots in presence of grazing incident sound waves with low or high intensities. Different slot configurations are designed based on the generic non-locally reacting liner model adopted in the experimental investigation. Both 2-D and 3-D numerical simulations of only slit resonators are carried out. Numerical results indicate that the extra peak is a result of a resonance excited in the second cavity at specific frequency. Under high sound pressure level incoming waves, the basic characteristics of the acoustic performance remain. However, vortex shedding transpires at the resonances around both the slits and the drainage slot. Vorticity contours show that the connection of two coupled cavities decreases the strength of vortex shedding around the basic Helmholtz resonance due to a higher energy reflection. Meanwhile, the cell interaction significantly increases the vorticity magnitude near the extra resonant frequency. Finally, a semi-empirical model is derived to predict the extra attenuation peak frequency.

  14. Slotted Photonic Crystal Sensors

    Science.gov (United States)

    Scullion, Mark G.; Krauss, Thomas F.; Di Falco, Andrea

    2013-01-01

    Optical biosensors are increasingly being considered for lab-on-a-chip applications due to their benefits such as small size, biocompatibility, passive behaviour and lack of the need for fluorescent labels. The light guiding mechanisms used by many of them results in poor overlap of the optical field with the target molecules, reducing the maximum sensitivity achievable. This review article presents a new platform for optical biosensors, namely slotted photonic crystals, which provide higher sensitivities due to their ability to confine, spatially and temporally, the optical mode peak within the analyte itself. Loss measurements showed values comparable to standard photonic crystals, confirming their ability to be used in real devices. A novel resonant coupler was designed, simulated, and experimentally tested, and was found to perform better than other solutions within the literature. Combining with cavities, microfluidics and biological functionalization allowed proof-of-principle demonstrations of protein binding to be carried out. Higher sensitivities were observed in smaller structures than possible with most competing devices reported in the literature. This body of work presents slotted photonic crystals as a realistic platform for complete on-chip biosensing; addressing key design, performance and application issues, whilst also opening up exciting new ideas for future study. PMID:23503295

  15. Slotted Photonic Crystal Sensors

    Directory of Open Access Journals (Sweden)

    Andrea Di Falco

    2013-03-01

    Full Text Available Optical biosensors are increasingly being considered for lab-on-a-chip applications due to their benefits such as small size, biocompatibility, passive behaviour and lack of the need for fluorescent labels. The light guiding mechanisms used by many of them results in poor overlap of the optical field with the target molecules, reducing the maximum sensitivity achievable. This review article presents a new platform for optical biosensors, namely slotted photonic crystals, which provide higher sensitivities due to their ability to confine, spatially and temporally, the optical mode peak within the analyte itself. Loss measurements showed values comparable to standard photonic crystals, confirming their ability to be used in real devices. A novel resonant coupler was designed, simulated, and experimentally tested, and was found to perform better than other solutions within the literature. Combining with cavities, microfluidics and biological functionalization allowed proof-of-principle demonstrations of protein binding to be carried out. Higher sensitivities were observed in smaller structures than possible with most competing devices reported in the literature. This body of work presents slotted photonic crystals as a realistic platform for complete on-chip biosensing; addressing key design, performance and application issues, whilst also opening up exciting new ideas for future study.

  16. Interpretation of chest radiographs with a high-resolution (2,000 x 2,000 x 12 bit) display

    International Nuclear Information System (INIS)

    Cox, G.G.

    1989-01-01

    This paper presents an evaluation of high-resolution (2K x 2Kx 12 bit) display for interpretation of chest radiographs. Three radiologists chose a total of 165 chest radiographs to ensure representation of nine signs: apical pleural scarring, chronic obstructive pulmonary disease, interstitial processes, atelectasis, pneumothorax, hilar mass, pleural effusion, pneumonia, and nodules. Each chest film was digitized to 4Kx 4Kx 12 bit and averaged to 2Kx 2Kx 12 bit and printed on a laser film printer. The 2K x 2K x 12-bit images were displayed and interactively windowed on a 2K x 2K x 12-bit high-resolution gray-scale cathode ray tube display. Six radiologists, none of whom participated in the case selection process, then interpreted a mixture of the screen film chest radiographs, the laser printed 2K chest radiographs, and the high resolution displayed 2K images

  17. Low profile frequency agile MIMO slot antenna with TCM characterization

    KAUST Repository

    Ghalib, Asim

    2017-06-07

    In this paper, a frequency reconfigurable multiple-input-multiple-output (MIMO) slot antenna is presented. The proposed design is low profile and compact with wide tunability range, covering several well-known frequency bands from 1800 MHz to 2450 MHz. The frequency reconfigurability is achieved by loading the annular slot with varactor diodes. The antenna system is also analyzed for MIMO performance metrics. Moreover, the effect of circular slot antenna on the chassis modes is also investigated using the theory of characteristic modes (TCM). The physical principle behind frequency reconfigurability is also investigated using TCM analysis. An interesting finding is observed using varactor diodes for frequency reconfigurability, that is the reactive impedance loading does not alter the modal significance (MS) plots but only aid in the input impedance matching at different frequency bands.

  18. Comparison of radiation dose, workflow, patient comfort and financial break-even of standard digital radiography and a novel biplanar low-dose X-ray system for upright full-length lower limb and whole spine radiography

    International Nuclear Information System (INIS)

    Dietrich, Tobias J.; Pfirrmann, Christian W.A.; Pankalla, Katja; Buck, Florian M.; Schwab, Alexander

    2013-01-01

    To compare the radiation dose, workflow, patient comfort, and financial break-even of a standard digital radiography and a biplanar low-dose X-ray system. A standard digital radiography system (Ysio, Siemens Healthcare, Erlangen, Germany) was compared with a biplanar X-ray unit (EOS, EOS imaging, Paris, France) consisting of two X-ray tubes and slot-scanning detectors, arranged at an angle of 90 allowing simultaneous vertical biplanar linear scanning in the upright patient position. We compared data of standing full-length lower limb radiographs and whole spine radiographs of both X-ray systems. Dose-area product was significantly lower for radiographs of the biplanar X-ray system than for the standard digital radiography system (e.g. whole spine radiographs; standard digital radiography system: 392.2 ± 231.7 cGy*cm 2 versus biplanar X-ray system: 158.4 ± 103.8 cGy*cm 2 ). The mean examination time was significantly shorter for biplanar radiographs compared with standard digital radiographs (e.g. whole spine radiographs: 449 s vs 248 s). Patients' comfort regarding noise was significantly higher for the standard digital radiography system. The financial break-even point was 2,602 radiographs/year for the standard digital radiography system compared with 4,077 radiographs/year for the biplanar X-ray unit. The biplanar X-ray unit reduces radiation exposure and increases subjective noise exposure to patients. The biplanar X-ray unit demands a higher number of examinations per year for the financial break-even point, despite the lower labour cost per examination due to the shorter examination time. (orig.)

  19. Ultrasound and supine chest radiograph in road traffic accident patients: a reliable and convenient way to diagnose pleural effusion

    International Nuclear Information System (INIS)

    Mumtaz, U.; Zahur, Z.; Raza, M.A.

    2017-01-01

    Background: Portable bed side ultrasound and supine chest radiograph of 80 traumatic patients excluding very clinically unstable patients who subsequently underwent CT scan chest was done for traumatic effusion showing that ultrasound had a higher sensitivity than CXR, 88.23% and 77.94%, respectively, and a similar specificity of 100% and 100%, respectively. Objective of the study is to compare the diagnostic accuracy of high resolution ultrasound and supine chest x-ray in detection of pleural effusion in road traffic accident patients keeping plain CT chest as gold standard. Methods: This study was conducted in PIMS and PAEC General Hospital, Islamabad from 1st January to 15th December 2015. The current study examined total of 80 trauma (blunt and penetrating) patients coming to emergency departments of both hospitals specifically those who had road traffic accident history. Their portable bed side ultrasound and supine chest radiograph were performed for assessing pleural effusion and subsequently CT scan chest was done for confirmation as it's a gold standard. Results: Using CT findings as gold standard the sensitivity, specificity, positive predictive value and negative predictive value was assessed for both ultrasonography and chest radiography and found to be 88.23%,100%, 100%, 40% and 77.94%, 100%, 100%, 55.55% respectively with diagnostic accuracy of ultrasound 90% as compared to 81.25% for supine chest x-rays when compared with gold standard. Conclusion: Ultrasound and chest x-ray can be used as a useful and suitable adjunct to CT in road traffic accident patients as these are easily available, non-invasive, no contrast required, can be performed on bed side and carries no or little radiation risk. (author)

  20. Delayed chest wall hematoma caused by progressive displacement of rib fractures after blunt trauma

    Directory of Open Access Journals (Sweden)

    Nobuhiro Sato

    2016-06-01

    Full Text Available Rib fracture is a common injury resulting from blunt thoracic trauma. Although hemothorax and pneumothorax are known delayed complications of rib fracture, delayed chest wall hematoma has rarely been reported. We discuss the case of an 81-year-old woman who was not undergoing antiplatelet or anticoagulant therapy who presented to our emergency department after a traffic injury. This patient had a nondisplaced rib fracture that went undetected on the initial computed tomography scan; the development of progressive displacement led to hemorrhagic shock due to delayed chest wall hematoma. The chest wall hematoma was effectively diagnosed and treated via contrast-enhanced computed tomography and angiographic embolization. This case highlights the possibility of this potential delayed complication from a common injury such as a rib fracture. Keywords: Angiography, Chest wall, Delayed complication, Rib fracture, Thoracic injury

  1. Rapid Prototyping — A Tool for Presenting 3-Dimensional Digital Models Produced by Terrestrial Laser Scanning

    Directory of Open Access Journals (Sweden)

    Juho-Pekka Virtanen

    2014-07-01

    Full Text Available Rapid prototyping has received considerable interest with the introduction of affordable rapid prototyping machines. These machines can be used to manufacture physical models from three-dimensional digital mesh models. In this paper, we compare the results obtained with a new, affordable, rapid prototyping machine, and a traditional professional machine. Two separate data sets are used for this, both of which were acquired using terrestrial laser scanning. Both of the machines were able to produce complex and highly detailed geometries in plastic material from models based on terrestrial laser scanning. The dimensional accuracies and detail levels of the machines were comparable, and the physical artifacts caused by the fused deposition modeling (FDM technique used in the rapid prototyping machines could be found in both models. The accuracy of terrestrial laser scanning exceeded the requirements for manufacturing physical models of large statues and building segments at a 1:40 scale.

  2. Wideband Circularly Polarized Printed Ring Slot Antenna for 5 GHz – 6 GHz

    Science.gov (United States)

    Nasrun Osman, Mohamed; Rahim, Mohamad Helmi A.; Jusoh, Muzammil; Sabapathy, Thennarasan; Rahim, Mohamad Kamal A.; Norlyana Azemi, Saidatul

    2018-03-01

    This paper presents the design of circularly polarized printed slot antenna operating at 5 – 6 GHz. The proposed antenna consists of L-shaped feedline on the top of structure and circular ring slot positioned at the ground plane underneath the substrate as a radiator. A radial and narrow slot in the ground plane provides coupling between the L-shaped feedline and circular ring slot. The circular polarization is realized by implementing the slits perturbation located diagonally to perturb the current flow on the slot structure. The antenna prototype is fabricated on FR4 substrate. The simulated and measured results are compared and analyzed to demonstrate the performance of the antenna. Good measured of simulated results are obtained at the targeted operating frequency. The simulated -10dB reflection coefficient bandwidths and axial ratio are 750 MHz and 165 MHz, respectively. The investigation on the affect of the important parameters towards the reflection coefficient and axial are also presented. The proposed antenna is highly potential to be used for wireless local area network (WLAN) and wireless power transfer (WPT).

  3. Modeling arbitrarily directed slots that are narrow both in width and depth with regard to the FDTD spatial cell

    Energy Technology Data Exchange (ETDEWEB)

    Riley, D.J.; Turner, C.D.

    1991-12-31

    The Hybrid Thin-Slot Algorithm (HTSA) integrates a transient integral-equation solution for an aperture in an infinite plane into a finite-difference time-domain (FDTD) code. The technique was introduced for linear apertures and was extended to include wall loss and lossy internal gaskets. A general implementation for arbitrary thin slots is briefly described here. The 3-D FDTD-code TSAR was selected for the implementation. The HTSA does not provide universal solutions to the narrow slot problem, but has merits appropriate for particular applications. The HTSA is restricted to planar slots, but can solve the important case that both the width and depth of the slot are narrow compared to the FDTD spatial cell. IN addition, the HTSA is not bound to the FDTD discrete spatial and time increments, and therefore, high-resolution solutions for the slot physics are possible. The implementation of the HTSA into TSAR is based upon a ``slot data file`` that includes the cell indices where the desired slots are exist within the FDTD mesh. For an HTSA-defined slot, the wall region local to the slot is shorted, and therefore, to change the slot`s topology simply requires altering the file to include the desired cells. 7 refs.

  4. [Digital luminescence radiography. A new method of study in thoracic diagnosis at the intensive care unit].

    Science.gov (United States)

    Witte, G; Pothmann, W; Bause, H; Nicolas, V; Schulte am Esch, J; Bücheler, E

    1989-02-01

    The digital luminescence-radiography (DLR) technique relies on a complete digitalization of the X-ray image. Luminescence crystals on the imaging plate serve as an energy reservoir following their exposure to ionized radiation from any conventional X-ray source. A Helium-Neon laser stimulates the electrons in their high energy bands and therefore will be dropped back emitting luminescence. This luminescence is digitized by the DLR-System thus delivering a complete digital image to the image processor for subsequent processing and evaluation. The processed digital image is then recorded on a conventional film or a monitor screen. More than 3000 chest examinations using DLR have been performed on intensive care unit (ICU) patients at the University Hospital Eppendorf following the first eleven months since the clinical introduction of this new technique. The positive aspects of DLR such as high-contrast resolution and optimal reproducibility were clinically evaluated under ICU conditions. It was shown that DLR greatly improves the quality of the chest X-rays of all ICU patients and offers the following advantages: reproducibility, lateral chest projection, no insufficient exposure, reduction of exposure dose, electronical post-processing and storage, quality preserving digital storage and copying.

  5. Development of a chest digital tomosynthesis R/F system and implementation of low-dose GPU-accelerated compressed sensing (CS) image reconstruction.

    Science.gov (United States)

    Choi, Sunghoon; Lee, Haenghwa; Lee, Donghoon; Choi, Seungyeon; Lee, Chang-Lae; Kwon, Woocheol; Shin, Jungwook; Seo, Chang-Woo; Kim, Hee-Joung

    2018-05-01

    This work describes the hardware and software developments of a prototype chest digital tomosynthesis (CDT) R/F system. The purpose of this study was to validate the developed system for its possible clinical application on low-dose chest tomosynthesis imaging. The prototype CDT R/F system was operated by carefully controlling the electromechanical subsystems through a synchronized interface. Once a command signal was delivered by the user, a tomosynthesis sweep started to acquire 81 projection views (PVs) in a limited angular range of ±20°. Among the full projection dataset of 81 images, several sets of 21 (quarter view) and 41 (half view) images with equally spaced angle steps were selected to represent a sparse view condition. GPU-accelerated and total-variation (TV) regularization strategy-based compressed sensing (CS) image reconstruction was implemented. The imaged objects were a flat-field using a copper filter to measure the noise power spectrum (NPS), a Catphan ® CTP682 quality assurance (QA) phantom to measure a task-based modulation transfer function (MTF T ask ) of three different cylinders' edge, and an anthropomorphic chest phantom with inserted lung nodules. The authors also verified the accelerated computing power over CPU programming by checking the elapsed time required for the CS method. The resultant absorbed and effective doses that were delivered to the chest phantom from two-view digital radiographic projections, helical computed tomography (CT), and the prototype CDT system were compared. The prototype CDT system was successfully operated, showing little geometric error with fast rise and fall times of R/F x-ray pulse less than 2 and 10 ms, respectively. The in-plane NPS presented essential symmetric patterns as predicted by the central slice theorem. The NPS images from 21 PVs were provided quite different pattern against 41 and 81 PVs due to aliased noise. The voxel variance values which summed all NPS intensities were inversely

  6. ATLAS Cold Leg Top Slot Break Analysis using RELAP5

    Energy Technology Data Exchange (ETDEWEB)

    Sung, Haejung; Lee, Sang Ik; Park, Ju-Hyun; Choi, Tong-Soo [KEPCO NF, Daejeon (Korea, Republic of)

    2016-10-15

    U.S. Nuclear Regulatory Commission (US-NRC) has been reviewing the design certification application for APR1400 submitted by Korea Electric Power Corporation (KEPCO). The main concern about cold leg top slot break is that cladding temperature might be increased by core uncover due to four loop seal reformation following flooding of safety injection water. An integral effect test for cold leg top slot break was performed by KAERI (Korea Atomic Energy Research Institute) using ATLAS (Advanced Thermal-Hydraulic Test Loop for Accident Simulation), which is a scaled down experimental facility for APR1400. In this study, RELAP5/MOD3.3/Patch04 is assessed by experimental result of ATLAS cold leg top slot break. Also, thermal hydraulic phenomena by four loop seals reformation is observed by RELAP5 result. The RELAP5/MOD3.3/Patch04 is assessed by the experimental result of ATLAS cold leg top slot break. The top slot break is described by offtake model, and the mass flow rate is fairly well estimated. The RELAP5 well predicts the correlation between general trend and four loop seal reformation. The pressure of the core region and the cladding temperature tends to increase during four loop seal reformation due to steam path blockage on four loop seals. It is presumed that the code cannot estimate two phase phenomena by loop seal clearing as same as experiments. In terms of cladding temperature, loop seal reformation due to loop seal elevation of APR1400 does not need to be the issue, since the void fraction at the active top core is maintained over 0.4.

  7. Indications for chest CT. Retrospective study of cases with normal chest CT

    International Nuclear Information System (INIS)

    Obata, Shiro

    1995-01-01

    The usefulness of computed tomography (CT) in thoracic radiology is now well appreciated, and the number of chest CTs has greatly increased. There are, however, many chest CT cases that are completely or almost completely normal. Indications for chest CT should be re-evaluated considering the cost and radiation exposure associated with the examination. Reviewing the reports of 4930 chest CT examinations performed in three hospitals during the period of two years, the author found 620 (12.6%) negative CT examinations. In 312 of the 620, the CT was requested because of 'abnormal shadow' on chest radiograph. When the same chest radiographs were re-evaluated by two radiologists, no abnormality was noted in 257 cases (82.4%). CT examinations were considered justified in only 55 cases (17.6%). There was a significant difference in the frequency of normal chest CT examinations between the university hospital and two other hospitals. The causes of false positive interpretation of chest radiographs were analyzed, and it was felt that fundamental knowledge necessary to interpret chest radiographs was lacking. The importance of close cooperation between clinicians and radiologists should be emphasized. (author)

  8. A Dual Lung Scan for the Evaluation of Pulmonary Function in Patients with Pulmonary Tuberculosis before and after Treatment

    International Nuclear Information System (INIS)

    Rhee, Chong Heon

    1967-01-01

    In 20 normal cases and 39 pulmonary tuberculosis cases, regional pulmonary arterial blood flow measurement and lung perfusion scans by 131 I-Macroaggregated albumin, lung inhalation scans by colloidal 198 Au and spirometries by respirometer were done at the Radiological Research Institute. The measured lung function tests were compared and the results were as the following: 1) The normal distribution of pulmonary blood flow was found to be 54.5±2.82% to the right lung and 45.5±2.39% to the left lung. The difference between the right and left pulmonary arterial blood flow was significant statistically (p 131 I-MAA in patients with pulmonary tuberculosis was as follows: a) In the pretreated minimal pulmonary tuberculosis, the decreased area of pulmonary arterial blood flow was corresponding to the chest roentgenogram, but the decrease of pulmonary arterial blood flow was more extensive than had been expected from the chest roentgenogram in the apparently healed minimal pulmonary tuberculosis. b) In the pretreated moderately advanced pulmonary tuberculosis, the decrease of pulmonary arterial blood flow to the diseased area was corresponding to the chest roentgenogram, but the decrease of pulmonary arterial blood flow was more extensive in the treated moderately advanced pulmonary tuberculosis as in the treated minimal pulmonary tuberculosis. c) Pulmonary arterial blood flow in the patients with far advanced pulmonary tuberculosis both before and after chemotherapy were almost similar to the chest roentgenogram. Especially the decrease of pulmonary arterial blood flow to the cavity was usually greater than had been expected from the chest roentgenogram. 3) Lung inhalation scan by colloidal 198 Au in patients with pulmonary tuberculosis was as follows: a) In the minimal pulmonary tuberculosis, lung inhalation scan showed almost similar decrease of radioactivity corresponding to the chest roentgenogram. b) In the moderately advanced pulmonary tuberculosis the decrease

  9. Fleet Planning Decision-Making: Two-Stage Optimization with Slot Purchase

    Directory of Open Access Journals (Sweden)

    Lay Eng Teoh

    2016-01-01

    Full Text Available Essentially, strategic fleet planning is vital for airlines to yield a higher profit margin while providing a desired service frequency to meet stochastic demand. In contrast to most studies that did not consider slot purchase which would affect the service frequency determination of airlines, this paper proposes a novel approach to solve the fleet planning problem subject to various operational constraints. A two-stage fleet planning model is formulated in which the first stage selects the individual operating route that requires slot purchase for network expansions while the second stage, in the form of probabilistic dynamic programming model, determines the quantity and type of aircraft (with the corresponding service frequency to meet the demand profitably. By analyzing an illustrative case study (with 38 international routes, the results show that the incorporation of slot purchase in fleet planning is beneficial to airlines in achieving economic and social sustainability. The developed model is practically viable for airlines not only to provide a better service quality (via a higher service frequency to meet more demand but also to obtain a higher revenue and profit margin, by making an optimal slot purchase and fleet planning decision throughout the long-term planning horizon.

  10. American College of Chest Physicians

    Science.gov (United States)

    ... Foundation Participate in the e-Community Get Social Career Connection Publications CHEST Journal CHEST SEEK Guidelines & Consensus Statements CHEST Physician CHEST NewsBrief Coding for Chest Medicine Tobacco Dependence Toolkit (3rd Ed.) Mobile Websites and Apps CHEST Journal ...

  11. Organ dose evaluation for CT scans based on in-phantom measurements

    International Nuclear Information System (INIS)

    Liu Haikuan; Zhuo Weihai; Chen Bo; Yi Yanling; Li Dehong

    2009-01-01

    Objective: To explore the organ doses and their distributions in different projections of CT scans. Methods: The CT values were measured and the linear absorption coefficients were derived for the main organs of the anthropomorphic phantom to compare with the normal values of human beings. The radiophotoluminescent glass dosimeters were set into various tissues or organs of the phantom for mimic measurements of the organ doses undergoing the head, chest, abdomen and pelvis CT scans, respectively. Results: The tissue equivalence of the phantom used in this study was good. The brain had the largest organ dose undergoing the head CT scan. The organ doses in thyroid, breast, lung and oesophagus were relatively large in performing the chest CT scan, while the liver, stomach, colon and lung had relatively hrge organ doses in abdomen CT practice. The doses in bone surface and colon exceeded by 50 mGy in a single pelvis CT scan. Conclusions: The organ doses and their distributions largely vary with different projections of CT scans. The organ doses of colon, bone marrow,gonads and bladder are fairly large in performing pelvis CT scan, which should be paid attention in the practice. (authors)

  12. Evaluation of digital dental models obtained from dental cone-beam computed tomography scan of alginate impressions

    OpenAIRE

    Jiang, Tingting; Lee, Sang-Mi; Hou, Yanan; Chang, Xin; Hwang, Hyeon-Shik

    2016-01-01

    Objective To investigate the dimensional accuracy of digital dental models obtained from the dental cone-beam computed tomography (CBCT) scan of alginate impressions according to the time elapse when the impressions are stored under ambient conditions. Methods Alginate impressions were obtained from 20 adults using 3 different alginate materials, 2 traditional alginate materials (Alginoplast and Cavex Impressional) and 1 extended-pour alginate material (Cavex ColorChange). The impressions wer...

  13. Extremely High-Birefringent Asymmetric Slotted-Core Photonic Crystal Fiber in THz Regime

    DEFF Research Database (Denmark)

    Islam, Raonaqul; Habib, Selim; Hasanuzzaman, G.K.M.

    2015-01-01

    We present a thorough numerical analysis of a highly birefringent slotted porous-core circular photonic crystal fiber (PCF) for terahertz (THz) wave guidance. The slot shaped air-holes break the symmetry of the porous-core which offers a very high birefringence whereas the compact geometry of the...

  14. Metal membrane with dimer slots as a universal polarizer

    DEFF Research Database (Denmark)

    Zhukovsky, Sergei; Zalkovskij, Maksim; Malureanu, Radu

    2014-01-01

    In this work, we show theoretically and confirm experimentally that thin metal membranes patterned with an array of slot dimers (or their Babinet analogue with metal rods) can function as a versatile spectral and polarization filter. We present a detailed covariant multipole theory for the electr......In this work, we show theoretically and confirm experimentally that thin metal membranes patterned with an array of slot dimers (or their Babinet analogue with metal rods) can function as a versatile spectral and polarization filter. We present a detailed covariant multipole theory...

  15. [Stress fractures of the ribs with acute thoracic pain in a young woman, diagnosed by the bone scan].

    Science.gov (United States)

    Georgitzikis, Athanasios; Siopi, Dimitra; Doumas, Argyrios; Mitka, Ekaterini; Antoniadis, Antonios

    2010-01-01

    We report the unusual case of a 29 -year old woman with emotional instability who presented with acute onset chest pain after severe chronic cough. The chest X-ray and the serological tests were normal but the CT scanning, and the bone scanning revealed multiple bilateral rib stress fractures, caused by severe coughing and physical activity and worsened by the patient's emotional instability.

  16. Chest MRI

    Science.gov (United States)

    ... resonance imaging - chest; NMR - chest; MRI of the thorax; Thoracic MRI Patient Instructions ... Gotway MB, Panse PM, Gruden JF, Elicker BM. Thoracic radiology. In: Broaddus VC, Mason RJ, Ernst JD, et ...

  17. TU-F-CAMPUS-I-01: Investigation of the Effective Dose From Bolus Tracking Acquisitions at Different Anatomical Locations in the Chest for CT

    Energy Technology Data Exchange (ETDEWEB)

    Nowik, P; Bujila, R; Merzan, D [Dept. of Medical Physics, Karolinska University Hospital, Stockholm (Sweden)

    2015-06-15

    Purpose: Stationary table acquisitions (Bolus tracking) in X-ray Computed Tomography (CT) can Result in dose length products (DLP) comparable to spiral scans. It is today unclear whether or not the effective dose (E) for Bolus Tracking can be approximated using target region specific conversion factors (E/DLP). The purpose of this study was to investigate how E depends on the anatomical location of the Bolus Tracking in relation to Chest CT scans with the same DLP. Methods: Effective doses were approximated for the ICRP 110 adult Reference Male (AM) and adult Reference Female (FM) computational voxel phantoms using software for CT dose approximations (pre-simulated MC data). The effective dose was first approximated for a Chest CT scan using spiral technique and a CTDIvol (32 cm) of 6 mGy. The effective dose from the spiral scan was then compared to E approximated for contiguous Bolus Tracking acquisitions (1 cm separation), with a total collimation of 1 cm, over different locations of the chest of the voxel phantoms. The number of rotations used for the Bolus Tracking acquisitions was adjusted to yield the same DLP (32 cm) as the spiral scan. Results: Depending on the anatomical location of the Bolus Tracking, E ranged by factors of 1.3 to 6.8 for the AM phantom and 1.4 to 3.3 for the AF phantom, compared to the effective dose of the spiral scans. The greatest E for the Bolus Tracking acquisitions was observed for anatomical locations coinciding with breast tissue. This can be expected as breast tissue has a high tissue weighting factor in the calculation of E. Conclusion: For Chest CT scans, the effective dose from Bolus Tracking is highly dependent on the anatomical location where the scan is administered and will not always accurately be represented using target region specific conversion factors.

  18. 14 CFR 93.218 - Slots for transborder service to and from Canada.

    Science.gov (United States)

    2010-01-01

    ... Canada. 93.218 Section 93.218 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF... service to and from Canada. (a) Except as otherwise provided in this subpart, international slots...'Hare in the Winter season. (c) Any modification to the slot base by the Government of Canada or the...

  19. Chest Pain: First Aid

    Science.gov (United States)

    First aid Chest pain: First aid Chest pain: First aid By Mayo Clinic Staff Causes of chest pain can vary from minor problems, such as indigestion ... 26, 2018 Original article: http://www.mayoclinic.org/first-aid/first-aid-chest-pain/basics/ART-20056705 . Mayo ...

  20. Severe Chest Wall Toxicity From Cryoablation in the Setting of Prior Stereotactic Ablative Radiotherapy.

    Science.gov (United States)

    Chaudhuri, Aadel A; Binkley, Michael S; Aggarwal, Sonya; Qian, Yushen; Carter, Justin N; Shah, Rajesh; Loo, Billy W

    2016-02-02

    We present the case of a 42-year-old woman with metastatic synovial sarcoma of parotid origin, treated definitively with chemoradiation, who subsequently developed oligometastatic disease limited to the lungs. She underwent multiple left and right lung wedge resections and left lower lobectomy, followed by right lower lobe stereotactic ablative radiotherapy (SABR), 54 Gy in three fractions to a right lower lobe lesion abutting the chest wall. Two years later, she was treated with cryoablation for a separate right upper lobe nodule abutting the chest wall. Two months later, she presented with acute shortness of breath, pleuritic chest pain, decreased peripheral blood O2 saturation, and productive cough. A computed tomography (CT) scan demonstrated severe chest wall necrosis in the area of recent cryoablation that, in retrospect, also received a significant radiation dose from her prior SABR. This case demonstrates that clinicians should exercise caution in using cryoablation when treating lung tumors abutting a previously irradiated chest wall. Note: Drs. Loo and Shah contributed equally as co-senior authors.

  1. A piezoelectric ultrasonic linear micromotor using a slotted stator.

    Science.gov (United States)

    Yun, Cheol-Ho; Watson, Brett; Friend, James; Yeo, Leslie

    2010-08-01

    A novel ultrasonic micro linear motor that uses 1st longitudinal and 2nd bending modes, derived from a bartype stator with a rectangular slot cut through the stator length, has been proposed and designed for end-effect devices of microrobotics and bio-medical applications. The slot structure plays an important role in the motor design, and can be used not only to tune the resonance frequency of the two vibration modes but also to reduce the undesirable longitudinal coupling displacement caused by bending vibration at the end of the stator. By using finite element analysis, the optimal slot dimension to improve the driving tip motion was determined, resulting in the improvement of the motor performance. The trial linear motor, with a weight of 1.6 g, gave a maximum driving velocity of 1.12 m/s and a maximum driving force of 3.4 N. A maximum mechanical output power of 1.1 W was obtained at force of 1.63 N and velocity of 0.68 m/s. The output mechanical power per unit weight was 688 W/kg.

  2. Seawave Slot-Cone Generator

    DEFF Research Database (Denmark)

    Vicinanza, Diego; Margheritini, Lucia; Contestabile, Pasquale

    2009-01-01

    This paper discusses a new type of Wave Energy Converter (WEC) named Seawave Slot-Cone Generator (SSG). The SSG is a WEC of the overtopping type. The structure consists of a number of reservoirs one on the top of each others above the mean water level in which the water of incoming waves is store...... on sloping walls constituting the structure. The research is intended to be of direct use to engineers analyzing design and stability of this peculiar kind of coastal structure....

  3. Pediatric chest imaging. Chest imaging in infants and children. 2. rev. ed.

    Energy Technology Data Exchange (ETDEWEB)

    Lucaya, Javier [Vall d' Hebron Hospitals, Barcelona (Spain). Dept. of Pediatric Radiology and Inst. of Diagnostic Imaging; Hospital Quiron, Barcelona (Spain). Dept. of Radiology; Strife, Janet L. (eds.) [Cincinnati Univ. Coll. of Medicine, Cincinnati, OH (United States). Dept. of Radiology Cincinnati Children' s Hospital Medical Center

    2008-07-01

    Imaging of the pediatric chest continues to evolve rapidly. All chapters in this 2nd edition of Pediatric Chest Imaging have been extensively updated, with additional disease-specific information and numerous new illustrations. The book thus presents the state of the art in the diagnosis of pediatric chest disorders, highlighting the role played by advanced technology. As the conventional features of most of these disorders are extremely well known, special attention is devoted to the technical aspects of the modern imaging modalities, their indications, and the diagnostic information that they supply. Individual chapters focus on chest ultrasound, nuclear medicine imaging, high-resolution chest CT, helical CT, and pediatric cardiac CT and pediatric cardiacMRI. Others are directed towards specific disorders, including congenital malformations of the chest, chest tumors, pulmonary infection, trauma, the lung in systemic diseases, the pediatric airway, foreign bodies, the thymus, and the chest wall. Without exception, the authors of this book are internationally known specialists with great expertise in the field. This book will serve as a handy, superbly illustrated reference for all who routinely image children, as well as for those who need access to information on how best to image them. (orig.)

  4. Ka-Band Slot-Microstrip-Covered and Waveguide-Cavity-Backed Monopulse Antenna Array

    Directory of Open Access Journals (Sweden)

    Li-Ming Si

    2014-01-01

    Full Text Available A slot-microstrip-covered and waveguide-cavity-backed monopulse antenna array is proposed for high-resolution tracking applications at Ka-band. The monopulse antenna array is designed with a microstrip with 2×32 slots, a waveguide cavity, and a waveguide monopulse comparator, to make the structure simple, reduce the feeding network loss, and increase the frequency bandwidth. The 2×32 slot-microstrip elements are formed by a metal clad dielectric substrate and slots etched in the metal using the standard printed circuit board (PCB process with dimensions of 230 mm  ×  10 mm. The proposed monopulse antenna array not only maintains the advantages of the traditional waveguide slot antenna array, but also has the characteristics of wide bandwidth, high consistence, easy of fabrication, and low cost. From the measured results, it exhibits good monopulse characteristics, including the following: the maximum gains of sum pattern are greater than 24 dB, the 3 dB beamwidth of sum pattern is about 2.2 degrees, the sidelobe levels of the sum pattern are less than −18 dB, and the null depths of the difference pattern are less than −25 dB within the operating bandwidth between 33.65 GHz and 34.35 GHz for VSWR ≤ 2.

  5. Evaluation of radiation doses delivered in different chest CT protocols

    International Nuclear Information System (INIS)

    Gorycki, Tomasz; Lasek, Iwona; Kamiński, Kamil; Studniarek, Michał

    2014-01-01

    There are differences in the reference diagnostic levels for the computed tomography (CT) of the chest as cited in different literature sources. The doses are expressed either in weighted CT dose index (CTDI VOL ) used to express the dose per slice, dose-length product (DLP), and effective dose (E). The purpose of this study was to assess the radiation dose used in Low Dose Computer Tomography (LDCT) of the chest in comparison with routine chest CT examinations as well as to compare doses delivered in low dose chest CT with chest X-ray doses. CTDI VOL and DLP doses were taken to analysis from routine CT chest examinations (64 MDCT TK LIGHT SPEED GE Medical System) performed in 202 adult patients with FBP reconstruction: 51 low dose, 106 helical, 20 angio CT, and 25 high resolution CT protocols, as well as 19 helical protocols with iterative ASIR reconstruction. The analysis of chest X-ray doses was made on the basis of reports from 44 examinations. Mean values of CTDI VOL and DLP were, respectively: 2.1 mGy and 85.1 mGy·cm, for low dose, 9.7 mGy and 392.3 mGy·cm for helical, 18.2 mGy and 813.9 mGy·cm for angio CT, 2.3 mGy and 64.4 mGy·cm for high resolution CT, 8.9 mGy. and 317.6 mGy·cm for helical ASIR protocols. Significantly lower CTDI VOL and DLP values were observed for low dose and high resolution CT versus the remaining CT protocols; doses delivered in CT ASIR protocols were also lower (80–81%). The ratio between medial doses in low dose CT and chest X-ray was 11.56. Radiation dose in extended chest LDCT with parameters allowing for identification of mediastinal structures and adrenal glands is still much lower than that in standard CT protocols. Effective doses predicted for LDCT may exceed those used in chest X-ray examinations by a factor of 4 to 12, depending on LDCT scan parameters. Our results, as well as results from other authors, suggest a possibility of reducing the dose by means of iterative reconstruction. Efforts towards further dose

  6. Inadvertent chest tube insertion in congenital cystic adenomatoid malformation and congenital lobar emphysema-highlighting an important problem

    International Nuclear Information System (INIS)

    Prabhu, Shailesh M; Choudhury, Subhasis Roy; Solanki, Ravi S; Shetty, Gurucharan S; Agarwala, Surenderkumar

    2013-01-01

    Chest tube insertion in congenital cystic lung lesions is an important problem in children with acute respiratory distress having a cystic lucent lesion on chest radiograph. To evaluate the imaging findings and complications in cases of congenital cystic lung lesions with chest tube insertion and suggest the role of appropriate imaging for management of these patients. Chest radiographs and CT scans of children with congenital cystic lung lesions who had inadvertent chest tube insertion preoperatively were retrospectively reviewed for imaging appearances and complications. Fifteen patients comprising 10 cases of congenital cystic adenomatoid malformation (CCAM) and 5 cases of congenital lobar emphysema (CLE) were included. Majority of the cases were infants. CCAM was misdiagnosed as complicated pneumatocele (n = 5) and pneumothorax (n = 5), while CLE was misdiagnosed as tension pneumothorax (n = 5) on the chest radiograph findings. Final diagnosis was made on CT and operative findings with histopathology. Complications noted were pneumothorax, hydropneumothorax, and infection in cases of CCAM, and change in imaging appearance and pneumothorax in cases of CLE. Chest tube insertion in congenital cystic lesions increases the rate of associated complications. Chest CT has a definite role in early diagnosis and deciding appropriate management in these cases

  7. Inflammation and Rupture of a Congenital Pericardial Cyst Manifesting Itself as an Acute Chest Pain Syndrome.

    Science.gov (United States)

    Aertker, Robert A; Cheong, Benjamin Y C; Lufschanowski, Roberto

    2016-12-01

    We present the case of a 63-year-old woman with a remote history of supraventricular tachycardia and hyperlipidemia, who presented with recurrent episodes of acute-onset chest pain. An electrocardiogram showed no evidence of acute coronary syndrome. A chest radiograph revealed a prominent right-sided heart border. A suspected congenital pericardial cyst was identified on a computed tomographic chest scan, and stranding was noted around the cyst. The patient was treated with nonsteroidal anti-inflammatory drugs, and the pain initially abated. Another flare-up was treated similarly. Cardiac magnetic resonance imaging was then performed after symptoms had resolved, and no evidence of the cyst was seen. The suspected cause of the patient's chest pain was acute inflammation of a congenital pericardial cyst with subsequent rupture and resolution of symptoms.

  8. TU-CD-207-05: A Novel Digital Tomosynthesis System Using Orthogonal Scanning Technique: A Feasibility Study

    Energy Technology Data Exchange (ETDEWEB)

    Kim, J [Samsung Medical Center, Seoul (Korea, Republic of); Park, C [University of Florida Health Science Center, Gainesville, FL (United States); Kauweloa, K [UTHSCSA, San Antonio, TX (United States); Chung, Y; Han, Y [Sungkyunkwan University School of Medicine, Seoul (Korea, Republic of); Ju, S [Samsung Medical Center, Seoul, Seoul (Korea, Republic of)

    2015-06-15

    Purpose: As an alternative to full tomographic imaging technique such as cone-beam computed tomography (CBCT), there is growing interest to adopt digital tomosynthesis (DTS) for the use of diagnostic as well as therapeutic applications. The aim of this study is to propose a new DTS system using novel orthogonal scanning technique, which can provide superior image quality DTS images compared to the conventional DTS scanning system. Methods: Unlike conventional DTS scanning system, the proposed DTS is reconstructed with two sets of orthogonal patient scans. 1) X-ray projections that are acquired along transverse trajectory and 2) an additional sets of X-ray projections acquired along the vertical direction at the mid angle of the previous transverse scan. To reconstruct DTS, we have used modified filtered backprojection technique to account for the different scanning directions of each projection set. We have evaluated the performance of our method using numerical planning CT data of liver cancer patient and a physical pelvis phantom experiment. The results were compared with conventional DTS techniques with single transverse and vertical scanning. Results: The experiments on both numerical simulation as well as physical experiment showed that the resolution as well as contrast of anatomical structures was much clearer using our method. Specifically, the image quality comparing with transversely scanned DTS showed that the edge and contrast of anatomical structures along Left-Right (LR) directions was comparable however, considerable discrepancy and enhancement could be observed along Superior-Inferior (SI) direction using our method. The opposite was observed when vertically scanned DTS was compared. Conclusion: In this study, we propose a novel DTS system using orthogonal scanning technique. The results indicated that the image quality of our novel DTS system was superior compared to conventional DTS system. This makes our DTS system potentially useful in various

  9. Synchronous-digitization for video rate polarization modulated beam scanning second harmonic generation microscopy

    Science.gov (United States)

    Sullivan, Shane Z.; DeWalt, Emma L.; Schmitt, Paul D.; Muir, Ryan D.; Simpson, Garth J.

    2015-03-01

    Fast beam-scanning non-linear optical microscopy, coupled with fast (8 MHz) polarization modulation and analytical modeling have enabled simultaneous nonlinear optical Stokes ellipsometry (NOSE) and linear Stokes ellipsometry imaging at video rate (15 Hz). NOSE enables recovery of the complex-valued Jones tensor that describes the polarization-dependent observables, in contrast to polarimetry, in which the polarization stated of the exciting beam is recorded. Each data acquisition consists of 30 images (10 for each detector, with three detectors operating in parallel), each of which corresponds to polarization-dependent results. Processing of this image set by linear fitting contracts down each set of 10 images to a set of 5 parameters for each detector in second harmonic generation (SHG) and three parameters for the transmittance of the fundamental laser beam. Using these parameters, it is possible to recover the Jones tensor elements of the sample at video rate. Video rate imaging is enabled by performing synchronous digitization (SD), in which a PCIe digital oscilloscope card is synchronized to the laser (the laser is the master clock.) Fast polarization modulation was achieved by modulating an electro-optic modulator synchronously with the laser and digitizer, with a simple sine-wave at 1/10th the period of the laser, producing a repeating pattern of 10 polarization states. This approach was validated using Z-cut quartz, and NOSE microscopy was performed for micro-crystals of naproxen.

  10. Diagnostic performance of digital breast tomosynthesis with a wide scan angle compared to full-field digital mammography for the detection and characterization of microcalcifications

    International Nuclear Information System (INIS)

    Clauser, Paola; Nagl, Georg; Helbich, Thomas H.; Pinker-Domenig, Katja; Weber, Michael; Kapetas, Panagiotis; Bernathova, Maria; Baltzer, Pascal A.T.

    2016-01-01

    Highlights: • Wide scan-angle DBT alone shows a high detection rate for microcalcifications. • DBT and FFDM can characterize microcalcifications at a comparable level. • Characterization is influenced by reader and by lesion type (benign vs malignant). • DBT might be used as a stand-alone technique for the assessment of microcalcifications. - Abstract: Objectives: To assess the diagnostic performance of digital breast tomosynthesis (DBT), with a wide scan-angle, compared to full-field digital mammography (FFDM), for the detection and characterization of microcalcifications. Methods: IRB approval was obtained for this retrospective study. We selected 150 FFDM and DBT (50 benign and 50 malignant histologically verified microcalcifications, 50 cases classified as BI-RADS 1). Four radiologists evaluated, in separate sessions and blinded to patients’ history and histology, the presence of microcalcifications. Cases with microcalcifications were assessed for visibility, characteristics, and grade of suspicion using BI-RADS categories. Detection rate and diagnostic performance were calculated. Visibility, lesions’ characteristics and reading time were analysed. Results: Detection rate and visibility were good for both FFDM and DBT, without intra-reader differences (P = 0.510). Inter-reader differences were detected (P < 0.018). Only two lesions were not detected by any reader on either FFDM or DBT. Diagnostic performance with DBT was as good as that of FFDM, but a significant inter-reader difference was found (P = 0.041). High inter-reader variability in the use of the descriptors was found. Reading time for DBT was almost twice that for FFDM (44 and 25 s, respectively). Conclusion: Wide scan-angle DBT enabled the detection and characterization of microcalcifications with no significant differences from FFDM. Inter-reader variability was seen.

  11. Diagnostic performance of digital breast tomosynthesis with a wide scan angle compared to full-field digital mammography for the detection and characterization of microcalcifications

    Energy Technology Data Exchange (ETDEWEB)

    Clauser, Paola, E-mail: paola.clauser@meduniwien.ac.at [Department of Biomedical Imaging and Image-Guided Therapy, Division of Molecular and Gender Imaging, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna (Austria); Nagl, Georg [Department for Radiology and Interventional Radiology, Landesklinikum Horn, Spitalgasse 10, 3580 Horn (Austria); Helbich, Thomas H., E-mail: thomas.helbich@meduniwien.ac.at [Department of Biomedical Imaging and Image-Guided Therapy, Division of Molecular and Gender Imaging, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna (Austria); Pinker-Domenig, Katja [Department of Biomedical Imaging and Image-Guided Therapy, Division of Molecular and Gender Imaging, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna (Austria); Weber, Michael [Department of Biomedical Imaging and Image-Guided Therapy, Division of General and Pediatric Radiology, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna (Austria); Kapetas, Panagiotis; Bernathova, Maria; Baltzer, Pascal A.T. [Department of Biomedical Imaging and Image-Guided Therapy, Division of Molecular and Gender Imaging, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna (Austria)

    2016-12-15

    Highlights: • Wide scan-angle DBT alone shows a high detection rate for microcalcifications. • DBT and FFDM can characterize microcalcifications at a comparable level. • Characterization is influenced by reader and by lesion type (benign vs malignant). • DBT might be used as a stand-alone technique for the assessment of microcalcifications. - Abstract: Objectives: To assess the diagnostic performance of digital breast tomosynthesis (DBT), with a wide scan-angle, compared to full-field digital mammography (FFDM), for the detection and characterization of microcalcifications. Methods: IRB approval was obtained for this retrospective study. We selected 150 FFDM and DBT (50 benign and 50 malignant histologically verified microcalcifications, 50 cases classified as BI-RADS 1). Four radiologists evaluated, in separate sessions and blinded to patients’ history and histology, the presence of microcalcifications. Cases with microcalcifications were assessed for visibility, characteristics, and grade of suspicion using BI-RADS categories. Detection rate and diagnostic performance were calculated. Visibility, lesions’ characteristics and reading time were analysed. Results: Detection rate and visibility were good for both FFDM and DBT, without intra-reader differences (P = 0.510). Inter-reader differences were detected (P < 0.018). Only two lesions were not detected by any reader on either FFDM or DBT. Diagnostic performance with DBT was as good as that of FFDM, but a significant inter-reader difference was found (P = 0.041). High inter-reader variability in the use of the descriptors was found. Reading time for DBT was almost twice that for FFDM (44 and 25 s, respectively). Conclusion: Wide scan-angle DBT enabled the detection and characterization of microcalcifications with no significant differences from FFDM. Inter-reader variability was seen.

  12. Usefulness of routine computed tomography in the evaluation of penetrating war injuries to the chest.

    Science.gov (United States)

    Darwish, Bassam; Mahfouz, Mohammad Z; Izzat, Mohammad Bashar

    2018-05-02

    This review was conducted to compare the contributions of chest X-ray (CXR) and computed tomography (CT) towards detecting intrathoracic damage in patients with penetrating war injuries to the chest and to determine whether identification of additional injuries by chest CT will have an impact on the choice of therapeutic interventions and clinical outcomes. We reviewed records of 449 patients (374 men, mean age 29.3 ± 14.8 years) who were admitted to our hospital with penetrating war injuries to the chest over a 7-year period. Collected data included mechanisms of injury, associated injuries, results of CXRs and chest CTs, methods of management, in-hospital stays, complications and mortalities. Immediate screening CXRs were obtained in all patients not requiring emergent thoracotomies, of which 91.4% showed positive signs of injury. Chest CTs were performed at the discretion of the physicians in 49.4% of patients, and CXR-positive findings were confirmed in all cases, while revealing additional injuries in 11% of patients. Chest CT findings led to additional closed chest drainage in 5.6% of patients but had no impact on treatment strategy in 94.4% of scanned patients. Follow-up CXRs showed new positive findings in 22 patients, leading to additional closed chest drainage in 3 patients and delayed open thoracotomies in 7 other patients. CXRs continue as the primary diagnostic modality in the assessment of patients with penetrating war injuries to the chest. Chest CTs can be omitted in most patients, thus reducing CT imaging case-load substantially, while most clinically significant chest injuries remain sufficiently recognized.

  13. Enhancement of acousto-optical coupling in two-dimensional air-slot phoxonic crystal cavities by utilizing surface acoustic waves

    Energy Technology Data Exchange (ETDEWEB)

    Ma, Tian-Xue [Institute of Engineering Mechanics, Beijing Jiaotong University, Beijing 100044 (China); Wang, Yue-Sheng, E-mail: yswang@bjtu.edu.cn [Institute of Engineering Mechanics, Beijing Jiaotong University, Beijing 100044 (China); Zhang, Chuanzeng [Department of Civil Engineering, University of Siegen, D-57068 Siegen (Germany)

    2017-01-30

    A phoxonic crystal is a periodically patterned material that can simultaneously localize optical and acoustic modes. The acousto-optical coupling in two-dimensional air-slot phoxonic crystal cavities is investigated numerically. The photons can be well confined in the slot owing to the large electric field discontinuity at the air/dielectric interfaces. Besides, the surface acoustic modes lead to the localization of the phonons near the air-slot. The high overlap of the photonic and phononic cavity modes near the slot results in a significant enhancement of the moving interface effect, and thus strengthens the total acousto-optical interaction. The results of two cavities with different slot widths show that the coupling strength is dependent on the slot width. It is expected to achieve a strong acousto-optical/optomechanical coupling in air-slot phoxonic crystal structures by utilizing surface acoustic modes. - Highlights: • Two-dimensional air-slot phoxonic crystal cavities which can confine simultaneously optical and acoustic waves are proposed. • The acoustic and optical waves are highly confined near/in the air-slot. • The high overlap of the photonic and phononic cavity modes significantly enhances the moving interface effect. • Different factors which affect the acousto-optical coupling are discussed.

  14. Subtle pulmonary nodules: detection and identification with storage phosphor radiographs and conventional chest films

    International Nuclear Information System (INIS)

    Scheck, R.J.; Schaetzl, M.; Kandziora, C.; Panzer, M.; Rienmueller, R.

    1994-01-01

    To determine the value of digital storagephosphor radiography (SR) on the detection and identification of subtle lung nodules, postero-anterior (PA) and lateral (LAT) film-screen (FR) chest radiographs were compared with isodose SR images of 45 patients with metastatic malignancies. The SR postprocessing was done with a particular mode previously optimized for routine chest radiography. Pulmonary metastases were found in 34 patients and were proved or excluded by CT (n=28) or longterm follow-up FR (n=17). Chest images were divided into four regions for evaluation of image quality, number of lung nodules per region and marked pulmonary structures by receiver-operating characteristics (ROC) analysis (45 patients; 125 nodules; 2810 observations; five readers). Of the nodules selected for an ROC study 82% were 0.5-1.0 cm in diameter. Overall image quality was rated better for FR concerning lung fields (PA) and mediastinum/hilum (LAT). More lung-field nodules were detected on FR than on SR chest images. Use of FR was superior to SR in the general identification of nodules (PA chest), especially concerning intermediate and subtle abnormalities, whereas there was no significant difference for LAT chest images. Our results show, that currently FR still has advantages over SR in the detection and identification of subtle lung nodules in routine clinical radiography. (orig.)

  15. Computer-aided detection of lung nodules in digital chest radiographs

    International Nuclear Information System (INIS)

    Giger, M.L.; Doi, K.; MacMahon, H.M.

    1986-01-01

    The authors are developing an automated method to detect lung nodules by eliminating the ''camouflaging: effect of the lung background. In order to increase the conspicuity of the nodules, we created, from a single chest radiograph, two images: one in which the signal-to-noise ratio (S/N) of the nodule is maximized and another in which that S/N is suppressed. The difference between these two processed images was subjected to feature-extraction technique in order to isolate the nodules. The detection accuracy of the computer-aided detection scheme, as compared with unaided radiologists' performance, was determined using receiver operating characteristic curve analysis

  16. Detailed Study of Closed Stator Slots for a Direct-Driven Synchronous Permanent Magnet Linear Wave Energy Converter

    Directory of Open Access Journals (Sweden)

    Erik Lejerskog

    2014-01-01

    Full Text Available The aim of this paper is to analyze how a permanent magnet linear generator for wave power behaves when the stator slots are closed. The usual design of stator geometry is to use open slots to maintain a low magnetic leakage flux between the stator teeth. By doing this, harmonics are induced in the magnetic flux density in the air-gap due to slotting. The closed slots are designed to cause saturation, to keep the permeability low. This reduces the slot harmonics in the magnetic flux density, but will also increase the flux leakage between the stator teeth. An analytical model has been created to study the flux through the closed slots and the result compared with finite element simulations. The outcome shows a reduction of the cogging force and a reduction of the harmonics of the magnetic flux density in the air-gap. It also shows a small increase of the total magnetic flux entering the stator and an increased magnetic flux leakage through the closed slots.

  17. SCAN- a maintenance-free flowrate meter with direct digital read-out for computerised control applications in radiochemical plants

    Energy Technology Data Exchange (ETDEWEB)

    Shah, B V; Siddiqui, I A; Theyyunni, T K [Process Engineering and Systems Division, Bhabha Atomic Research Centre, Mumbai (India)

    1994-06-01

    In radiochemical plants, the choice of flowrate sensor is subject to stringent requirements of fail-safe design and freedom from maintenance. The SCAN remote digital direct indicating flowrate meter described in this paper was developed to meet the requirements of flowrate and transmitting it to the control room, and to a computerised control system. SCAN is designed on the principle that flowrate through an orifice is a function of the head of liquid acting upon it. SCAN consists of a small chamber which receives the flow, and discharges it through an orifice located in bottom. The level of liquid in the pot represents the flowrate of the input stream. SCAN has been developed into an accurate, rugged and practical device by refinements in the design of internals, and by introducing a special end-section which makes the calibration insensitive to location. An important feature of SCAN is that it is passive, maintenance free, fail-safe device and contains no moving parts. There is no liquid hold up in the SCAN when idle, which is a desirable feature for the radiochemical plant environment. (author). 3 figs., 2 tabs.

  18. Fast transient digitizer

    International Nuclear Information System (INIS)

    Villa, F.

    1982-01-01

    Method and apparatus for sequentially scanning a plurality of target elements with an electron scanning beam modulated in accordance with variations in a high-frequency analog signal to provide discrete analog signal samples representative of successive portions of the analog signal; coupling the discrete analog signal samples from each of the target elements to a different one of a plurality of high speed storage devices; converting the discrete analog signal samples to equivalent digital signals; and storing the digital signals in a digital memory unit for subsequent measurement or display

  19. Chest computed tomography

    DEFF Research Database (Denmark)

    Loeve, Martine; Krestin, Gabriel P.; Rosenfeld, Margaret

    2013-01-01

    are not suitable to study CF lung disease in young children. Chest computed tomography (CT) holds great promise for use as a sensitive surrogate endpoint in CF. A large body of evidence has been produced to validate the use of chest CT as primary endpoint to study CF lung disease. However, before chest CT can...

  20. Comparison of digital scanning and polyvinyl siloxane impression techniques by dental students: instructional efficiency and attitudes towards technology.

    Science.gov (United States)

    Marti, A M; Harris, B T; Metz, M J; Morton, D; Scarfe, W C; Metz, C J; Lin, W-S

    2017-08-01

    With increasing use of digital scanning with restorative procedures in the dental office, it becomes necessary that educational institutions adopt instructional methodology for introducing this technology together with conventional impression techniques. To compare the time differences between instructing dental students on digital scanning (DS) (LAVA C.O.S. digital impression system) and a conventional impression technique (CI) (polyvinyl siloxane), and to compare students' attitudes and beliefs towards both techniques. Volunteer sophomore dental students (n = 25) with no prior experience in clinical impressions were recruited and IRB consent obtained. Participants responded to a pre-and post-exposure questionnaire. Participants were instructed on the use of both DS and CI for a single tooth full coverage crown restoration using a consecutive sequence of video lecture, investigator-led demonstration and independent impression exercise. The time necessary for each step (minutes) was recorded. Statistical significance was calculated using dependent t-tests (time measurements) and 2-sample Mann-Whitney (questionnaire responses). The time spent teaching students was greater for DS than CI for video lecture (15.95 and 10.07 min, P = 0.0000), demonstration time (9.06 and 4.70 min, P = 0.0000) and impression time (18.17 and 8.59 min, P = 0.0000). Prior to the instruction and practice, students considered themselves more familiar with CI (3.96) than DS (1.96) (P = 0.0000). After the instruction and practice, participants reported CI technique proved significantly easier than expected (pre-instruction: 3.52 and post-instruction: 4.08, P = 0.002). However, overall participants' perception of ease of use for DS was not influenced by this instruction and practice experience (pre-instruction: 3.84 and post-instruction: 3.56, P = 0.106). Despite the results, 96% of participants expressed an expectation that DS will become their predominant impression technique during their