  1. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

    Science.gov (United States)

    Rowshanfarzad, P; Riis, H L; Zimmermann, S J; Ebert, M A

    2015-07-01

    In radiotherapy treatments, it is crucial to monitor the performance of linear accelerator (linac) components, including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt and the sag in leaf bank assembly owing to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with 5 ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of the image data. Nine Elekta AB (Stockholm, Sweden) linacs of different models and number of years in service were investigated. The average EPID sag was within 2 mm for all tested linacs. Some machines showed >1-mm gantry sag. Changes in the SDD values were within 1.3 cm. EPID skewness and tilt values were <1° in all machines. The maximum sag in multileaf collimator leaf bank assemblies was around 1 mm. A meaningful correlation was found between the age of the linacs and their mechanical performance. Conclusions and Advances in knowledge: The method and software developed in this study provide a simple tool for effective investigation of the behaviour of Elekta linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Elekta machines.

  2. SU-E-P-45: An Analytical Formula for Deriving Mechanical Iso-Center of Rotational Gantry Treatment Unit Rotational Gantry Treatment Unit

    International Nuclear Information System (INIS)

    Ding, X; Bues, M

    2015-01-01

    Purpose: To present an analytical formula for deriving mechanical isocenter (MIC) of a rotational gantry treatment unit. The input data to the formula is obtained by a custom-made device. The formula has been implemented and used in an operational proton therapy facility since 2005. Methods: The custom made device consisted of 3 mutually perpendicular dial indicators and 5 clinometers, to obtain displacement data and gantry angle data simultaneously. During measurement, a steel sphere was affixed to the patient couch, and the device was attached to the snout rotating with the gantry. The displacement data and angle data were obtained simultaneously at angular increments of less than 1 degree. The analytical formula took the displacement and angle as input and derived the positions of dial indicator tips (DIT) position in room-fixed coordinate system. The formula derivation presupposes trigonometry and 3-dimentional coordinate transformations. Due to the symmetry properties of the defining equations, the DIT position can be solved for analytically without using mathematical approximations. We define the mean of all points in the DIT trajectory as the MIC. The formula was implemented in computer code, which has been employed during acceptance test, commissioning, as well as routine QA practice in an operational proton facility since 2005. Results: It took one minute for the custom-made device to acquire the measurement data for a full gantry rotation. The DIT trajectory and MIS are instantaneously available after the measurement. The MIC Result agrees well with vendor’s Result, which came from a different measurement setup, as well as different data analysis algorithm. Conclusion: An analytical formula for deriving mechanical isocenter was developed and validated. The formula is considered to be absolutely accurate mathematically. Be analyzing measured data of radial displacements as function of gantry angle, the formula calculates the MI position in room

  3. Proposal of cancer therapy system without rotating gantry

    International Nuclear Information System (INIS)

    Kodaira, Masanobu

    2002-01-01

    Beam therapy is one of useful methods for cancer therapy. Many results in National Institute of Radiological Sciences (NIRS) show many abilities of beam therapy for cancer therapy. In Japan, several beam therapy facilities are constructed or under construction. If its construction budget becomes to be smaller, beam therapy may be used as the general cancer therapy. But in the present beam therapy facilities, the budget of its construction is very large. One of the reasons of big budget is the construction of the big buildings equipped with thick shielding walls. Most of space of the facilities with thick shielding walls is devoted to the treatment equipments such as rotating gantries and beam transport lines. This proposal is that using oblique beam line and rotating treatment bed, multi-portal irradiation is realized without rotating gantry. At the same time, we designed adequate beam lines to minimize the total facilities. (author)

  4. Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device.

    Science.gov (United States)

    Rowshanfarzad, Pejman; Häring, Peter; Riis, Hans L; Zimmermann, Sune J; Ebert, Martin A

    2015-01-01

    In radiotherapy treatments, it is crucial to monitor the performance of linac components including gantry, collimation system, and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in leaf bank assembly due to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with five ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of image data. Three Siemens linacs were investigated. The average EPID sag was within 1 mm for all tested linacs. Two machines showed >1 mm gantry sag. Changes in the SDD values were within 7.5 mm. EPID skewness and tilt values were <1° in all machines. The maximum sag in leaf bank assembly was <1 mm. The method and software developed in this study provide a simple tool for effective investigation of the behavior of Siemens linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Siemens machines.

  5. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation

    DEFF Research Database (Denmark)

    Rowshanfarzad, P; Lynggaard Riis, Hans; Zimmermann, S J

    2015-01-01

    OBJECTIVE: In radiotherapy treatments, it is crucial to monitor the performance of linear accelerator (linac) components, including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested...... collimator leaf bank assemblies was around 1 mm. A meaningful correlation was found between the age of the linacs and their mechanical performance. Conclusions and Advances in knowledge: The method and software developed in this study provide a simple tool for effective investigation of the behaviour...

  6. Design and Test Results of Superconducting Magnet for Heavy-Ion Rotating Gantry

    Science.gov (United States)

    Takayama, S.; Koyanagi, K.; Miyazaki, H.; Takami, S.; Orikasa, T.; Ishii, Y.; Kurusu, T.; Iwata, Y.; Noda, K.; Obana, T.; Suzuki, K.; Ogitsu, T.; Amemiya, N.

    2017-07-01

    Heavy-ion radiotherapy has a high curative effect in cancer treatment and also can reduce the burden on patients. These advantages have been generally recognized. Furthermore, a rotating gantry can irradiate a tumor with ions from any direction without changing the position of the patient. This can reduce the physical dose on normal cells, and is thus commonly used in proton radiotherapy. However, because of the high magnetic rigidity of carbon ions, the weight of the rotating gantry for heavy-ion therapy is about three-times heavier than those used for proton cancer therapy, according to our estimation. To overcome this issue, we developed a small and lightweight rotating gantry in collaboration with the National Institute of Radiological Sciences (NIRS). The compact rotating gantry was composed of ten low-temperature superconducting (LTS) magnets that were designed from the viewpoint of beam optics. These LTS magnets have a surface-winding coil-structure and provide both dipole and quadrupole fields. The maximum dipole and quadrupole magnetic field of the magnets were 2.88 T and 9.3 T/m, respectively. The rotating gantry was installed at NIRS, and beam commissioning is in progress to achieve the required beam quality. In the three years since 2013, in a project supported by the Ministry of Economy, Trade and Industry (METI) and the Japan Agency for Medical Research and Development (AMED), we have been developing high-temperature superconducting (HTS) magnets with the aim of a further size reduction of the rotating gantry. To develop fundamental technologies for designing and fabricating HTS magnets, a model magnet was manufactured. The model magnet was composed of 24 saddle-shaped HTS coils and generated a magnetic field of 1.2 T. In the presentation, recent progress in this research will be reported.

  7. Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device

    Directory of Open Access Journals (Sweden)

    Rowshanfarzad P

    2015-11-01

    Full Text Available Pejman Rowshanfarzad,1 Peter Häring,2 Hans L Riis,3 Sune J Zimmermann,3 Martin A Ebert1,4 1School of Physics, The University of Western Australia, Crawley, WA, Australia; 2German Cancer Research Center (DKFZ, Medical Physics in Radiation Oncology, Heidelberg, Germany; 3Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark; 4Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, Australia Background: In radiotherapy treatments, it is crucial to monitor the performance of linac components including gantry, collimation system, and electronic portal imaging device (EPID during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. Methods: The EPID sag, gantry sag, changes in source-to-detector distance (SDD, EPID and collimator skewness, EPID tilt, and the sag in leaf bank assembly due to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with five ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of image data. Three Siemens linacs were investigated. Results: The average EPID sag was within 1 mm for all tested linacs. Two machines showed >1 mm gantry sag. Changes in the SDD values were within 7.5 mm. EPID skewness and tilt values were <1° in all machines. The maximum sag in leaf bank assembly was <1 mm. Conclusion: The method and software developed in this study provide a simple tool for effective investigation of the behavior of Siemens linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Siemens machines. Keywords: linac, Siemens, arc, sag, EPID, gantry

  8. Measurements of isocenter path characteristics of the gantry rotation axis with a smartphone application

    International Nuclear Information System (INIS)

    Schiefer, H.; Peters, S.; Plasswilm, L.; Ingulfsen, N.; Kluckert, J.

    2015-01-01

    Purpose: For stereotactic radiosurgery, the AAPM Report No. 54 [AAPM Task Group 42 (AAPM, 1995)] requires the overall stability of the isocenter (couch, gantry, and collimator) to be within a 1 mm radius. In reality, a rotating system has no rigid axis and thus no isocenter point which is fixed in space. As a consequence, the isocenter concept is reviewed here. It is the aim to develop a measurement method following the revised definitions. Methods: The mechanical isocenter is defined here by the point which rotates on the shortest path in the room coordinate system. The path is labeled as “isocenter path.” Its center of gravity is assumed to be the mechanical isocenter. Following this definition, an image-based and radiation-free measurement method was developed. Multiple marker pairs in a plane perpendicular to the assumed gantry rotation axis of a linear accelerator are imaged with a smartphone application from several rotation angles. Each marker pair represents an independent measuring system. The room coordinates of the isocenter path and the mechanical isocenter are calculated based on the marker coordinates. The presented measurement method is by this means strictly focused on the mechanical isocenter. Results: The measurement result is available virtually immediately following completion of measurement. When 12 independent measurement systems are evaluated, the standard deviations of the isocenter path points and mechanical isocenter coordinates are 0.02 and 0.002 mm, respectively. Conclusions: The measurement is highly accurate, time efficient, and simple to adapt. It is therefore suitable for regular checks of the mechanical isocenter characteristics of the gantry and collimator rotation axis. When the isocenter path is reproducible and its extent is in the range of the needed geometrical accuracy, it should be taken into account in the planning process. This is especially true for stereotactic treatments and radiosurgery

  9. Measurements of isocenter path characteristics of the gantry rotation axis with a smartphone application

    Energy Technology Data Exchange (ETDEWEB)

    Schiefer, H., E-mail: johann.schiefer@kssg.ch; Peters, S.; Plasswilm, L. [Klinik für Radio-Onkologie, Kantonsspital St.Gallen, Rorschacherstrasse 107, St.Gallen CH-9007 (Switzerland); Ingulfsen, N.; Kluckert, J. [Kantonsschule am Burggraben St.Gallen, Burggraben 21, St.Gallen CH-9000 (Switzerland)

    2015-03-15

    Purpose: For stereotactic radiosurgery, the AAPM Report No. 54 [AAPM Task Group 42 (AAPM, 1995)] requires the overall stability of the isocenter (couch, gantry, and collimator) to be within a 1 mm radius. In reality, a rotating system has no rigid axis and thus no isocenter point which is fixed in space. As a consequence, the isocenter concept is reviewed here. It is the aim to develop a measurement method following the revised definitions. Methods: The mechanical isocenter is defined here by the point which rotates on the shortest path in the room coordinate system. The path is labeled as “isocenter path.” Its center of gravity is assumed to be the mechanical isocenter. Following this definition, an image-based and radiation-free measurement method was developed. Multiple marker pairs in a plane perpendicular to the assumed gantry rotation axis of a linear accelerator are imaged with a smartphone application from several rotation angles. Each marker pair represents an independent measuring system. The room coordinates of the isocenter path and the mechanical isocenter are calculated based on the marker coordinates. The presented measurement method is by this means strictly focused on the mechanical isocenter. Results: The measurement result is available virtually immediately following completion of measurement. When 12 independent measurement systems are evaluated, the standard deviations of the isocenter path points and mechanical isocenter coordinates are 0.02 and 0.002 mm, respectively. Conclusions: The measurement is highly accurate, time efficient, and simple to adapt. It is therefore suitable for regular checks of the mechanical isocenter characteristics of the gantry and collimator rotation axis. When the isocenter path is reproducible and its extent is in the range of the needed geometrical accuracy, it should be taken into account in the planning process. This is especially true for stereotactic treatments and radiosurgery.

  10. Four-dimensional cone beam CT with adaptive gantry rotation and adaptive data sampling

    International Nuclear Information System (INIS)

    Lu Jun; Guerrero, Thomas M.; Munro, Peter; Jeung, Andrew; Chi, P.-C. M.; Balter, Peter; Zhu, X. Ronald; Mohan, Radhe; Pan Tinsu

    2007-01-01

    We have developed a new four-dimensional cone beam CT (4D-CBCT) on a Varian image-guided radiation therapy system, which has radiation therapy treatment and cone beam CT imaging capabilities. We adapted the speed of gantry rotation time of the CBCT to the average breath cycle of the patient to maintain the same level of image quality and adjusted the data sampling frequency to keep a similar level of radiation exposure to the patient. Our design utilized the real-time positioning and monitoring system to record the respiratory signal of the patient during the acquisition of the CBCT data. We used the full-fan bowtie filter during data acquisition, acquired the projection data over 200 deg of gantry rotation, and reconstructed the images with a half-scan cone beam reconstruction. The scan time for a 200-deg gantry rotation per patient ranged from 3.3 to 6.6 min for the average breath cycle of 3-6 s. The radiation dose of the 4D-CBCT was about 1-2 times the radiation dose of the 4D-CT on a multislice CT scanner. We evaluated the 4D-CBCT in scanning, data processing and image quality with phantom studies. We demonstrated the clinical applicability of the 4D-CBCT and compared the 4D-CBCT and the 4D-CT scans in four patient studies. The contrast-to-noise ratio of the 4D-CT was 2.8-3.5 times of the contrast-to-noise ratio of the 4D-CBCT in the four patient studies

  11. Measuring the wobble of radiation field centers during gantry rotation and collimator movement on a linear accelerator

    International Nuclear Information System (INIS)

    Du, Weiliang; Gao, Song

    2011-01-01

    Purpose: The isocenter accuracy of a linear accelerator is often assessed with star-shot films. This approach is limited in its ability to quantify three dimensional wobble of radiation field centers (RFCs). The authors report a Winston-Lutz based method to measure the 3D wobble of RFCs during gantry rotation, collimator rotation, and collimator field size change. Methods: A stationary ball-bearing phantom was imaged using multileaf collimator-shaped radiation fields at various gantry angles, collimator angles, and field sizes. The center of the ball-bearing served as a reference point, to which all RFCs were localized using a computer algorithm with subpixel accuracy. Then, the gantry rotation isocenter and the collimator rotation axis were derived from the coordinates of these RFCs. Finally, the deviation or wobble of the individual RFC from the derived isocenter or rotation axis was quantified. Results: The results showed that the RFCs were stable as the field size of the multileaf collimator was varied. The wobble of RFCs depended on the gantry angle and the collimator angle and was reproducible, indicating that the mechanical imperfections of the linac were mostly systematic and quantifiable. It was found that the 3D wobble of RFCs during gantry rotation was reduced after compensating for a constant misalignment of the multileaf collimator. Conclusions: The 3D wobble of RFCs can be measured with submillimeter precision using the proposed method. This method provides a useful tool for checking and adjusting the radiation isocenter tightness of a linac.

  12. Measuring the wobble of radiation field centers during gantry rotation and collimator movement on a linear accelerator

    Energy Technology Data Exchange (ETDEWEB)

    Du, Weiliang; Gao, Song [Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas, 77030 (United States)

    2011-08-15

    Purpose: The isocenter accuracy of a linear accelerator is often assessed with star-shot films. This approach is limited in its ability to quantify three dimensional wobble of radiation field centers (RFCs). The authors report a Winston-Lutz based method to measure the 3D wobble of RFCs during gantry rotation, collimator rotation, and collimator field size change. Methods: A stationary ball-bearing phantom was imaged using multileaf collimator-shaped radiation fields at various gantry angles, collimator angles, and field sizes. The center of the ball-bearing served as a reference point, to which all RFCs were localized using a computer algorithm with subpixel accuracy. Then, the gantry rotation isocenter and the collimator rotation axis were derived from the coordinates of these RFCs. Finally, the deviation or wobble of the individual RFC from the derived isocenter or rotation axis was quantified. Results: The results showed that the RFCs were stable as the field size of the multileaf collimator was varied. The wobble of RFCs depended on the gantry angle and the collimator angle and was reproducible, indicating that the mechanical imperfections of the linac were mostly systematic and quantifiable. It was found that the 3D wobble of RFCs during gantry rotation was reduced after compensating for a constant misalignment of the multileaf collimator. Conclusions: The 3D wobble of RFCs can be measured with submillimeter precision using the proposed method. This method provides a useful tool for checking and adjusting the radiation isocenter tightness of a linac.

  13. Development of a superconducting rotating-gantry for heavy-ion therapy

    International Nuclear Information System (INIS)

    Iwata, Y.; Noda, K.; Murakami, T.; Shirai, T.; Furukawa, T.; Fujita, T.; Mori, S.; Mizushima, K.; Shouda, K.; Fujimoto, T.; Ogitsu, T.; Obana, T.; Amemiya, N.; Orikasa, T.; Takami, S.; Takayama, S.

    2013-01-01

    An isocentric superconducting rotating-gantry for heavy-ion therapy is being developed. This rotating gantry can transport heavy ions having 430 MeV/u to an isocenter with irradiation angles of over ±180°, and is further capable of performing fast raster-scanning irradiation. For the magnets, combined-function superconducting-magnets are to be employed. All of the superconducting magnets have been designed, and their magnetic-field distributions were calculated by using a 3D electromagnetic-field solver. With the calculated magnetic-fields, beam-tracking simulations were performed to evaluate the field quality of the superconducting magnets. Beam profiles as well as phase-space distributions at the isocenter, as calculated by simulations, agreed well with those of linear beam-optics calculations, proving validity of the final design for the magnets. Three superconducting magnets were manufactured, and are being tested. In this paper, results of tracking simulations as well as test results for the superconducting magnets are presented

  14. Effect of MLC leaf position, collimator rotation angle, and gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Bai, Sen; Li, Guangjun; Wang, Maojie; Jiang, Qinfeng; Zhang, Yingjie [State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu, Sichuan (China); Wei, Yuquan, E-mail: yuquawei@vip.sina.com [State Key Laboratory of Biotherapy and Cancer Center, West China Hospital, Sichuan University, Chengdu, Sichuan (China)

    2013-07-01

    The purpose of this study was to investigate the effect of multileaf collimator (MLC) leaf position, collimator rotation angle, and accelerator gantry rotation angle errors on intensity-modulated radiotherapy plans for nasopharyngeal carcinoma. To compare dosimetric differences between the simulating plans and the clinical plans with evaluation parameters, 6 patients with nasopharyngeal carcinoma were selected for simulation of systematic and random MLC leaf position errors, collimator rotation angle errors, and accelerator gantry rotation angle errors. There was a high sensitivity to dose distribution for systematic MLC leaf position errors in response to field size. When the systematic MLC position errors were 0.5, 1, and 2 mm, respectively, the maximum values of the mean dose deviation, observed in parotid glands, were 4.63%, 8.69%, and 18.32%, respectively. The dosimetric effect was comparatively small for systematic MLC shift errors. For random MLC errors up to 2 mm and collimator and gantry rotation angle errors up to 0.5°, the dosimetric effect was negligible. We suggest that quality control be regularly conducted for MLC leaves, so as to ensure that systematic MLC leaf position errors are within 0.5 mm. Because the dosimetric effect of 0.5° collimator and gantry rotation angle errors is negligible, it can be concluded that setting a proper threshold for allowed errors of collimator and gantry rotation angle may increase treatment efficacy and reduce treatment time.

  15. Rotational micro-CT using a clinical C-arm angiography gantry

    International Nuclear Information System (INIS)

    Patel, V.; Hoffmann, K. R.; Ionita, C. N.; Keleshis, C.; Bednarek, D. R.; Rudin, S.

    2008-01-01

    Rotational angiography (RA) gantries are used routinely to acquire sequences of projection images of patients from which 3D renderings of vascular structures are generated using Feldkamp cone-beam reconstruction algorithms. However, these systems have limited resolution ( 10 lp/mm) but to date have relied either on rotating object imaging or small bore geometry for small animal imaging, and thus are not used for clinical imaging. The authors report here the development and use of a 3D rotational micro-angiography (RMA) system created by mounting a micro-angiographic fluoroscope (MAF) [35 μm pixel, resolution >10 lp/mm, field of view (FOV)=3.6 cm] on a standard clinical FPD-based RA gantry (Infinix, Model RTP12303J-G9E, Toshiba Medical Systems Corp., Tustin, CA). RA image sequences are obtained using the MAF and reconstructed. To eliminate artifacts due to image truncation, lower-dose (compared to MAF acquisition) full-FOV (FFOV) FPD RA sequences (194 μm pixel, FOV=20 cm) were also obtained to complete the missing data. The RA gantry was calibrated using a helical bead phantom. To ensure high-quality high-resolution reconstruction, the high-resolution images from the MAF were aligned spatially with the lower-dose FPD images, and the pixel values in the FPD image data were scaled to match those of the MAF. Images of a rabbit with a coronary stent placed in an artery in the Circle of Willis were obtained and reconstructed. The MAF images appear well aligned with the FPD images (average correlation coefficient before and after alignment: 0.65 and 0.97, respectively) Greater details without any visible truncation artifacts are seen in 3D RMA (MAF-FPD) images than in those of the FPD alone. The FWHM of line profiles of stent struts (100 μm diameter) are approximately 192±21 and 313±38 μm for the 3D RMA and FPD data, respectively. In addition, for the dual-acquisition 3D RMA, FFOV FPD data need not be of the highest quality, and thus may be acquired at lower dose

  16. Quantifying the gantry sag on linear accelerators and introducing an MLC-based compensation strategy

    Energy Technology Data Exchange (ETDEWEB)

    Du Weiliang; Gao Song; Wang Xiaochun; Kudchadker, Rajat J. [Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, Texas 77030 (United States)

    2012-04-15

    Purpose: Gantry sag is one of the well-known sources of mechanical imperfections that compromise the spatial accuracy of radiation dose delivery. The objectives of this study were to quantify the gantry sag on multiple linear accelerators (linacs), to investigate a multileaf collimator (MLC)-based strategy to compensate for gantry sag, and to verify the gantry sag and its compensation with film measurements. Methods: The authors used the Winston-Lutz method to measure gantry sag on three Varian linacs. A ball bearing phantom was imaged with megavolt radiation fields at 10 deg. gantry angle intervals. The images recorded with an electronic portal imaging device were analyzed to derive the radiation isocenter and the gantry sag, that is, the superior-inferior wobble of the radiation field center, as a function of the gantry angle. The authors then attempted to compensate for the gantry sag by applying a gantry angle-specific correction to the MLC leaf positions. The gantry sag and its compensation were independently verified using film measurements. Results: Gantry sag was reproducible over a six-month measurement period. The maximum gantry sag was found to vary from 0.7 to 1.0 mm, depending on the linac and the collimator angle. The radiation field center moved inferiorly (i.e., away from the gantry) when the gantry was rotated from 0 deg. to 180 deg. After the MLC leaf position compensation was applied at 90 deg. collimator angle, the maximum gantry sag was reduced to <0.2 mm. The film measurements at gantry angles of 0 deg. and 180 deg. verified the inferior shift of the radiation fields and the effectiveness of MLC compensation. Conclusions: The results indicate that gantry sag on a linac can be quantitatively measured using a simple phantom and an electronic portal imaging device. Reduction of gantry sag is feasible by applying a gantry angle-specific correction to MLC leaf positions at 90 deg. collimator angle.

  17. Beam-transport study of an isocentric rotating ion gantry with minimum number of quadrupoles

    International Nuclear Information System (INIS)

    Pavlovic, Marius; Griesmayer, Erich; Seemann, Rolf

    2005-01-01

    A beam-transport study of an isocentric gantry for ion therapy is presented. The gantry is designed with the number of quadrupoles down to the theoretical minimum, which is the feature published for the first time in this paper. This feature has been achieved without compromising the ion-optical functions of the beam-transport system that is capable of handling non-symmetric beams (beams with different emittances in vertical and horizontal plane), pencil-beam scanning, double-achromatic optics and beam-size control. Ion-optical properties of the beam-transport system are described, discussed and illustrated by computer simulations performed by the TRANSPORT-code

  18. Gantry cranes

    NARCIS (Netherlands)

    Luttekes, E.; Rijsenbrij, J.C.

    2003-01-01

    Portal crane (1) for loading and/or unloading a ship having a multiplicity of bays for containers (10) comprising a gantry (2) with travelling gear (5) for supporting the gantry on a quay (13), a boom supported by the gantry made up of a horizontal beam (6) and a jib (7) that during loading and/or

  19. Gantry for computed tomography

    International Nuclear Information System (INIS)

    Kelman, A.L.; Peterson, T.E.

    1981-01-01

    A novel design of gantry for use in computed tomography is described in detail. In the new gantry, curved tracks are mounted to the laterally spaced apart sides of the frame which rotates and carries the detector and X-ray source. This permits the frame to be tilted either side of vertical enabling angular slices of body layers to be viewed and allows simplification of the algorithm which the computer uses for image reconstruction. The tracks are supported on rollers which carry the substantial weight. Explicit engineering details are presented especially of the ball bearing races used in the rotation. (U.K.)

  20. Gantry for computed tomography

    International Nuclear Information System (INIS)

    Kelman, A.L.; O'Dell, W.R.; Brook, R.F.; Hein, P.W.; Brandt, R.T.

    1981-01-01

    A novel design of gantry for use in computed tomography is described in detail. In the new gantry, curved tracks are mounted to the laterally spaced apart sides of the frame which rotates and carries the detector and X-ray source. This permits the frame to be tilted either side of vertical enabling angular slices of body layers to be viewed and allows simplification of the algorithm which the computer uses for image reconstruction. A failsafe, solenoid brake is described which can lock the shaft against rotation. The gantry also contains a hoist mechanism which aids maintenance of the heavy X-ray tube and/or detector arrays. Explicit engineering details are presented. (U.K.)

  1. MO-FG-202-04: Gantry-Resolved Linac QA for VMAT: A Comprehensive and Efficient System Using An Electronic Portal Imaging Device

    Energy Technology Data Exchange (ETDEWEB)

    Zwan, B J [Central Coast Cancer Centre, Gosford, NSW (Australia); University of Newcastle, Newcastle, NSW (Australia); Barnes, M; Greer, P B [University of Newcastle, Newcastle, NSW (Australia); Calvary Mater Hospital, Newcastle, NSW (Australia); Hindmarsh, J; Seymour, E [Central Coast Cancer Centre, Gosford, NSW (Australia); O’Connor, D J [University of Newcastle, Newcastle, NSW (Australia); Keall, P J [University of Sydney, Camperdown, NSW (Australia)

    2016-06-15

    Purpose: To automate gantry-resolved linear accelerator (linac) quality assurance (QA) for volumetric modulated arc therapy (VMAT) using an electronic portal imaging device (EPID). Methods: A QA system for VMAT was developed that uses an EPID, frame-grabber assembly and in-house developed image processing software. The system relies solely on the analysis of EPID image frames acquired without the presence of a phantom. Images were acquired at 8.41 frames per second using a frame grabber and ancillary acquisition computer. Each image frame was tagged with a gantry angle from the linac’s on-board gantry angle encoder. Arc-dynamic QA plans were designed to assess the performance of each individual linac component during VMAT. By analysing each image frame acquired during the QA deliveries the following eight machine performance characteristics were measured as a function of gantry angle: MLC positional accuracy, MLC speed constancy, MLC acceleration constancy, MLC-gantry synchronisation, beam profile constancy, dose rate constancy, gantry speed constancy, dose-gantry angle synchronisation and mechanical sag. All tests were performed on a Varian iX linear accelerator equipped with a 120 leaf Millennium MLC and an aS1000 EPID (Varian Medical Systems, Palo Alto, CA, USA). Results: Machine performance parameters were measured as a function of gantry angle using EPID imaging and compared to machine log files and the treatment plan. Data acquisition is currently underway at 3 centres, incorporating 7 treatment units, at 2 weekly measurement intervals. Conclusion: The proposed system can be applied for streamlined linac QA and commissioning for VMAT. The set of test plans developed can be used to assess the performance of each individual components of the treatment machine during VMAT deliveries as a function of gantry angle. The methodology does not require the setup of any additional phantom or measurement equipment and the analysis is fully automated to allow for

  2. MO-FG-202-04: Gantry-Resolved Linac QA for VMAT: A Comprehensive and Efficient System Using An Electronic Portal Imaging Device

    International Nuclear Information System (INIS)

    Zwan, B J; Barnes, M; Greer, P B; Hindmarsh, J; Seymour, E; O’Connor, D J; Keall, P J

    2016-01-01

    Purpose: To automate gantry-resolved linear accelerator (linac) quality assurance (QA) for volumetric modulated arc therapy (VMAT) using an electronic portal imaging device (EPID). Methods: A QA system for VMAT was developed that uses an EPID, frame-grabber assembly and in-house developed image processing software. The system relies solely on the analysis of EPID image frames acquired without the presence of a phantom. Images were acquired at 8.41 frames per second using a frame grabber and ancillary acquisition computer. Each image frame was tagged with a gantry angle from the linac’s on-board gantry angle encoder. Arc-dynamic QA plans were designed to assess the performance of each individual linac component during VMAT. By analysing each image frame acquired during the QA deliveries the following eight machine performance characteristics were measured as a function of gantry angle: MLC positional accuracy, MLC speed constancy, MLC acceleration constancy, MLC-gantry synchronisation, beam profile constancy, dose rate constancy, gantry speed constancy, dose-gantry angle synchronisation and mechanical sag. All tests were performed on a Varian iX linear accelerator equipped with a 120 leaf Millennium MLC and an aS1000 EPID (Varian Medical Systems, Palo Alto, CA, USA). Results: Machine performance parameters were measured as a function of gantry angle using EPID imaging and compared to machine log files and the treatment plan. Data acquisition is currently underway at 3 centres, incorporating 7 treatment units, at 2 weekly measurement intervals. Conclusion: The proposed system can be applied for streamlined linac QA and commissioning for VMAT. The set of test plans developed can be used to assess the performance of each individual components of the treatment machine during VMAT deliveries as a function of gantry angle. The methodology does not require the setup of any additional phantom or measurement equipment and the analysis is fully automated to allow for

  3. Gantry for computed tomography

    International Nuclear Information System (INIS)

    Brandt, R.T.; Hein, P.W.

    1981-01-01

    A novel design of gantry for use in computed tomography is described in detail. In the new gantry, curved tracks are mounted to the laterally spaced apart sides of the frame which rotates and carries the detector and X-ray source. This permits the frame to be tilted either side of vertical enabling angular slices of body layers to be viewed and allows simplification of the algorithm which the computer uses for image reconstruction. The tracks are supported on rollers which carry the substantial weight. Explicit engineering details are presented. (U.K.)

  4. An ion-optical design study of a carbon-ion rotating gantry with a superconducting final bending magnet

    International Nuclear Information System (INIS)

    Bokor, J.; Pavlovič, M.

    2016-01-01

    Ion-optical designs of an isocentric ion gantry with a compact curved superconducting final bending magnet are presented. The gantry is designed for transporting carbon-therapy beams with nominal kinetic energy of 400 MeV/u, which corresponds to the penetration range of C"6"+ beam in water of about 28 cm. In contrast to other existing designs, we present a “hybrid” beam transport system containing a single superconducting element – the last bending magnet. All other elements are based on conventional warm technology. Ion-optical properties of such a hybrid system are investigated in case of transporting non-symmetric (i.e. different emittance patterns in the horizontal and vertical plane) beams. Different conditions for transporting the non-symmetric beams are analyzed aiming at finding the optimal, i.e. the most compact, gantry version. The final gantry layout is presented including a 2D parallel scanning. The ion-optical and scanning properties of the final gantry design are described, discussed and illustrated by computer simulations performed by WinAGILE.

  5. SU-E-T-195: Gantry Angle Dependency of MLC Leaf Position Error

    Energy Technology Data Exchange (ETDEWEB)

    Ju, S; Hong, C; Kim, M; Chung, K; Kim, J; Han, Y; Ahn, S; Chung, S; Shin, E; Shin, J; Kim, H; Kim, D; Choi, D [Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul (Korea, Republic of)

    2014-06-01

    Purpose: The aim of this study was to investigate the gantry angle dependency of the multileaf collimator (MLC) leaf position error. Methods: An automatic MLC quality assurance system (AutoMLCQA) was developed to evaluate the gantry angle dependency of the MLC leaf position error using an electronic portal imaging device (EPID). To eliminate the EPID position error due to gantry rotation, we designed a reference maker (RM) that could be inserted into the wedge mount. After setting up the EPID, a reference image was taken of the RM using an open field. Next, an EPID-based picket-fence test (PFT) was performed without the RM. These procedures were repeated at every 45° intervals of the gantry angle. A total of eight reference images and PFT image sets were analyzed using in-house software. The average MLC leaf position error was calculated at five pickets (-10, -5, 0, 5, and 10 cm) in accordance with general PFT guidelines using in-house software. This test was carried out for four linear accelerators. Results: The average MLC leaf position errors were within the set criterion of <1 mm (actual errors ranged from -0.7 to 0.8 mm) for all gantry angles, but significant gantry angle dependency was observed in all machines. The error was smaller at a gantry angle of 0° but increased toward the positive direction with gantry angle increments in the clockwise direction. The error reached a maximum value at a gantry angle of 90° and then gradually decreased until 180°. In the counter-clockwise rotation of the gantry, the same pattern of error was observed but the error increased in the negative direction. Conclusion: The AutoMLCQA system was useful to evaluate the MLC leaf position error for various gantry angles without the EPID position error. The Gantry angle dependency should be considered during MLC leaf position error analysis.

  6. Rotational IMRT techniques compared to fixed gantry IMRT and Tomotherapy: multi-institutional planning study for head-and-neck cases

    Directory of Open Access Journals (Sweden)

    Lutters Gerd

    2011-02-01

    Full Text Available Abstract Background Recent developments enable to deliver rotational IMRT with standard C-arm gantry based linear accelerators. This upcoming treatment technique was benchmarked in a multi-center treatment planning study against static gantry IMRT and rotational IMRT based on a ring gantry for a complex parotid gland sparing head-and-neck technique. Methods Treatment plans were created for 10 patients with head-and-neck tumours (oropharynx, hypopharynx, larynx using the following treatment planning systems (TPS for rotational IMRT: Monaco (ELEKTA VMAT solution, Eclipse (Varian RapidArc solution and HiArt for the helical tomotherapy (Tomotherapy. Planning of static gantry IMRT was performed with KonRad, Pinnacle and Panther DAO based on step&shoot IMRT delivery and Eclipse for sliding window IMRT. The prescribed doses for the high dose PTVs were 65.1Gy or 60.9Gy and for the low dose PTVs 55.8Gy or 52.5Gy dependend on resection status. Plan evaluation was based on target coverage, conformity and homogeneity, DVHs of OARs and the volume of normal tissue receiving more than 5Gy (V5Gy. Additionally, the cumulative monitor units (MUs and treatment times of the different technologies were compared. All evaluation parameters were averaged over all 10 patients for each technique and planning modality. Results Depending on IMRT technique and TPS, the mean CI values of all patients ranged from 1.17 to 2.82; and mean HI values varied from 0.05 to 0.10. The mean values of the median doses of the spared parotid were 26.5Gy for RapidArc and 23Gy for VMAT, 14.1Gy for Tomo. For fixed gantry techniques 21Gy was achieved for step&shoot+KonRad, 17.0Gy for step&shoot+Panther DAO, 23.3Gy for step&shoot+Pinnacle and 18.6Gy for sliding window. V5Gy values were lowest for the sliding window IMRT technique (3499 ccm and largest for RapidArc (5480 ccm. The lowest mean MU value of 408 was achieved by Panther DAO, compared to 1140 for sliding window IMRT. Conclusions All

  7. Rotational IMRT techniques compared to fixed gantry IMRT and Tomotherapy: multi-institutional planning study for head-and-neck cases

    International Nuclear Information System (INIS)

    Wiezorek, Tilo; Schubert, Kai; Wagner, Daniela; Wendt, Thomas G; Brachwitz, Tim; Georg, Dietmar; Blank, Eyck; Fotina, Irina; Habl, Gregor; Kretschmer, Matthias; Lutters, Gerd; Salz, Henning

    2011-01-01

    Recent developments enable to deliver rotational IMRT with standard C-arm gantry based linear accelerators. This upcoming treatment technique was benchmarked in a multi-center treatment planning study against static gantry IMRT and rotational IMRT based on a ring gantry for a complex parotid gland sparing head-and-neck technique. Treatment plans were created for 10 patients with head-and-neck tumours (oropharynx, hypopharynx, larynx) using the following treatment planning systems (TPS) for rotational IMRT: Monaco (ELEKTA VMAT solution), Eclipse (Varian RapidArc solution) and HiArt for the helical tomotherapy (Tomotherapy). Planning of static gantry IMRT was performed with KonRad, Pinnacle and Panther DAO based on step&shoot IMRT delivery and Eclipse for sliding window IMRT. The prescribed doses for the high dose PTVs were 65.1Gy or 60.9Gy and for the low dose PTVs 55.8Gy or 52.5Gy dependend on resection status. Plan evaluation was based on target coverage, conformity and homogeneity, DVHs of OARs and the volume of normal tissue receiving more than 5Gy (V 5Gy ). Additionally, the cumulative monitor units (MUs) and treatment times of the different technologies were compared. All evaluation parameters were averaged over all 10 patients for each technique and planning modality. Depending on IMRT technique and TPS, the mean CI values of all patients ranged from 1.17 to 2.82; and mean HI values varied from 0.05 to 0.10. The mean values of the median doses of the spared parotid were 26.5Gy for RapidArc and 23Gy for VMAT, 14.1Gy for Tomo. For fixed gantry techniques 21Gy was achieved for step&shoot+KonRad, 17.0Gy for step&shoot+Panther DAO, 23.3Gy for step&shoot+Pinnacle and 18.6Gy for sliding window. V 5Gy values were lowest for the sliding window IMRT technique (3499 ccm) and largest for RapidArc (5480 ccm). The lowest mean MU value of 408 was achieved by Panther DAO, compared to 1140 for sliding window IMRT. All IMRT delivery technologies with their associated TPS

  8. Computerized tomographic x-ray scanner system and gantry assembly

    International Nuclear Information System (INIS)

    Doyd, D.P.; Lanzara, G.

    1991-01-01

    This patent describes a scanner assembly. It comprises: a support head, a C-shaped gantry, means for supporting the gantry in the support head for rotating movement, an x-ray source mounted on one side of the gantry for independent movement with respect to a detector array along the gantry, the x-ray source projecting x-rays across the gantry, and a detector array mounted on the other side of a gantry for independent movement with respect to the x-ray source along the gantry, the detector array serving to receive the projected x-rays

  9. Impact of gantry rotation time on plan quality and dosimetric verification. Volumetric modulated arc therapy (VMAT) vs. intensity modulated radiotherapy (IMRT)

    Energy Technology Data Exchange (ETDEWEB)

    Pasler, Marlies; Wirtz, Holger; Lutterbach, Johannes [Gemeinschaftspraxis fuer Strahlentherapie Singen-Friedrichshafen, Singen (Germany)

    2011-12-15

    To compare plan quality criteria and dosimetric accuracy of step-and-shoot intensity-modulated radiotherapy (ss-IMRT) and volumetric modulated arc radiotherapy (VMAT) using two different gantry rotation times. This retrospective planning study based on 20 patients was comprised of 10 prostate cancer (PC) and 10 head and neck (HN) cancer cases. Each plan contained two target volumes: a primary planning target volume (PTV) and a boost volume. For each patient, one ss-IMRT plan and two VMAT plans at 90 s (VMAT90) and 120 s (VMAT120) per arc were generated with the Pinnacle {sup copyright} planning system. Two arcs were provided for the PTV plans and a single arc for boost volumes. Dosimetric verification of the plans was performed using a 2D ionization chamber array placed in a full scatter phantom. VMAT reduced delivery time and monitor units for both treatment sites compared to IMRT. VMAT120 vs. VMAT90 increased delivery time and monitor units in PC plans without improving plan quality. For HN cases, VMAT120 provided comparable organs at risk sparing and better target coverage and conformity than VMAT90. In the VMAT plan verification, an average of 97.1% of the detector points passed the 3 mm, 3% {gamma} criterion, while in IMRT verification it was 98.8%. VMAT90, VMAT120, and IMRT achieved comparable treatment plans. Slower gantry movement in VMAT120 plans only improves dosimetric quality for highly complex targets.

  10. Impact of gantry rotation time on plan quality and dosimetric verification. Volumetric modulated arc therapy (VMAT) vs. intensity modulated radiotherapy (IMRT)

    International Nuclear Information System (INIS)

    Pasler, Marlies; Wirtz, Holger; Lutterbach, Johannes

    2011-01-01

    To compare plan quality criteria and dosimetric accuracy of step-and-shoot intensity-modulated radiotherapy (ss-IMRT) and volumetric modulated arc radiotherapy (VMAT) using two different gantry rotation times. This retrospective planning study based on 20 patients was comprised of 10 prostate cancer (PC) and 10 head and neck (HN) cancer cases. Each plan contained two target volumes: a primary planning target volume (PTV) and a boost volume. For each patient, one ss-IMRT plan and two VMAT plans at 90 s (VMAT90) and 120 s (VMAT120) per arc were generated with the Pinnacle copyright planning system. Two arcs were provided for the PTV plans and a single arc for boost volumes. Dosimetric verification of the plans was performed using a 2D ionization chamber array placed in a full scatter phantom. VMAT reduced delivery time and monitor units for both treatment sites compared to IMRT. VMAT120 vs. VMAT90 increased delivery time and monitor units in PC plans without improving plan quality. For HN cases, VMAT120 provided comparable organs at risk sparing and better target coverage and conformity than VMAT90. In the VMAT plan verification, an average of 97.1% of the detector points passed the 3 mm, 3% γ criterion, while in IMRT verification it was 98.8%. VMAT90, VMAT120, and IMRT achieved comparable treatment plans. Slower gantry movement in VMAT120 plans only improves dosimetric quality for highly complex targets.

  11. SU-F-T-457: A Filmless Method for Measurement of Couch Translation Per Gantry Rotation and Couch Speed for Tomotherapy Using ArcCheck

    International Nuclear Information System (INIS)

    Yang, B; Wong, R; Geng, H; Lam, W; Cheung, K; Yu, S

    2016-01-01

    Purpose: To develop a filmless methodology based on an ArcCheck for QA measurement of the couch translation per gantry rotation and couch speed of a Tomotherapy unit. Methods: Two test plans recommended by TG148 were chosen for this study. A helical plan with 1 cm field size opened the leaves for 180 degrees at the 2nd, 7th and 12th of total 13 rotations and was used to verify if the couch travelled the expected distance per gantry rotation. The other test plan was a static plan with the gantry at 0°, 1cm field width and constant couch movement speed of 0.5mm/s. It was used for couch speed measurement. The ArcCheck was isocentrically set up and recorded movie files which took a snapshot exposure every 50ms. Due to the spiral pattern of diodes distribution, when one of the diodes of the ArcCheck located at the beam center, the dose profile as measured by the row of diodes which surrounded the center diode should have a symmetrical pattern. We divided the profile into left half A and right half B. Then a shape parameter was defined as S=Σ|(A−B)|/Σ(A+B). By searching the local minimum of S parameter, the beam center at different time could be located. The machine trajectory log data were also collected and analyzed for comparison. Results: The mean value of measured couch translation and couch speed by ArcCheck had less than 0.05% deviation from the planned values. For couch speed measurement, our result showed a mean value of 0.5002 with an uncertainty ±0.0031, which agreed very well with both the plan setting of 0.5 mm/s and the machine log data of 0.5005 mm/s. Conclusion: Couch translation measured using ArcCheck is accurate and comparable to the film-based measurement. This filmless method also provides a convenient and independent way for measuring couch speed.

  12. SU-F-T-457: A Filmless Method for Measurement of Couch Translation Per Gantry Rotation and Couch Speed for Tomotherapy Using ArcCheck

    Energy Technology Data Exchange (ETDEWEB)

    Yang, B; Wong, R; Geng, H; Lam, W; Cheung, K; Yu, S [Hong Kong Sanatorium & Hospital, Hong Kong, N/A, Hong Kong (China)

    2016-06-15

    Purpose: To develop a filmless methodology based on an ArcCheck for QA measurement of the couch translation per gantry rotation and couch speed of a Tomotherapy unit. Methods: Two test plans recommended by TG148 were chosen for this study. A helical plan with 1 cm field size opened the leaves for 180 degrees at the 2nd, 7th and 12th of total 13 rotations and was used to verify if the couch travelled the expected distance per gantry rotation. The other test plan was a static plan with the gantry at 0°, 1cm field width and constant couch movement speed of 0.5mm/s. It was used for couch speed measurement. The ArcCheck was isocentrically set up and recorded movie files which took a snapshot exposure every 50ms. Due to the spiral pattern of diodes distribution, when one of the diodes of the ArcCheck located at the beam center, the dose profile as measured by the row of diodes which surrounded the center diode should have a symmetrical pattern. We divided the profile into left half A and right half B. Then a shape parameter was defined as S=Σ|(A−B)|/Σ(A+B). By searching the local minimum of S parameter, the beam center at different time could be located. The machine trajectory log data were also collected and analyzed for comparison. Results: The mean value of measured couch translation and couch speed by ArcCheck had less than 0.05% deviation from the planned values. For couch speed measurement, our result showed a mean value of 0.5002 with an uncertainty ±0.0031, which agreed very well with both the plan setting of 0.5 mm/s and the machine log data of 0.5005 mm/s. Conclusion: Couch translation measured using ArcCheck is accurate and comparable to the film-based measurement. This filmless method also provides a convenient and independent way for measuring couch speed.

  13. Conditions for reliable time-resolved dosimetry of electronic portal imaging devices for fixed-gantry IMRT and VMAT

    International Nuclear Information System (INIS)

    Yeo, Inhwan Jason; Patyal, Baldev; Mandapaka, Anant; Jung, Jae Won; Yi, Byong Yong; Kim, Jong Oh

    2013-01-01

    Purpose: The continuous scanning mode of electronic portal imaging devices (EPID) that offers time-resolved information has been newly explored for verifying dynamic radiation deliveries. This study seeks to determine operating conditions (dose rate stability and time resolution) under which that mode can be used accurately for the time-resolved dosimetry of intensity-modulated radiation therapy (IMRT) beams.Methods: The authors have designed the following test beams with variable beam holdoffs and dose rate regulations: a 10 × 10 cm open beam to serve as a reference beam; a sliding window (SW) beam utilizing the motion of a pair of multileaf collimator (MLC) leaves outside the 10 × 10 cm jaw; a step and shoot (SS) beam to move the pair in step; a volumetric modulated arc therapy (VMAT) beam. The beams were designed in such a way that they all produce the same open beam output of 10 × 10 cm. Time-resolved ion chamber measurements at isocenter and time-resolved and integrating EPID measurements were performed for all beams. The time-resolved EPID measurements were evaluated through comparison with the ion chamber and integrating EPID measurements, as the latter are accepted procedures. For two-dimensional, time-resolved evaluation, a VMAT beam with an infield MLC travel was designed. Time-resolved EPID measurements and Monte Carlo calculations of such EPID dose images for this beam were performed and intercompared.Results: For IMRT beams (SW and SS), the authors found disagreement greater than 2%, caused by frame missing of the time-resolved mode. However, frame missing disappeared, yielding agreement better than 2%, when the dose rate of irradiation (and thus the frame acquisition rates) reached a stable and planned rate as the dose of irradiation was raised past certain thresholds (a minimum 12 s of irradiation per shoot used for SS IMRT). For VMAT, the authors found that dose rate does not affect the frame acquisition rate, thereby causing no frame missing

  14. Evaluation of computed tomography coronary angiography in patients with a high heart rate using 16-slice spiral computed tomography with 0.37-s gantry rotation time

    International Nuclear Information System (INIS)

    Zhang, Shi-Zheng; Hu, Xiu-Hua; Zhang, Qiao-Wei; Huang, Wen-Xin

    2005-01-01

    The aim of our study is to evaluate computed tomography (CT) coronary angiography in patients with a high heart rate using 16-slice spiral CT with 0.37-s gantry rotation time. We compare the image quality of patients whose heart rates were over 70 beats per minute (bpm) with that of patients whose heart rates were 70 bpm or less. Sixty patients with various heart rates underwent retrospectively ECG-gated multislice spiral CT (MSCT) coronary angiography. Two experienced observers who were blind to the heart rates of the patients evaluated all the MSCT coronary angiographic images and calculated the assessable segments. A total of 620 out of 891 (69.6%) segments were satisfactorily visualized. On average, 10.3 coronary artery segments per patient could be evaluated. In 36 patients whose heart rates were below 70 bpm [mean 62.2 bpm±5.32 (standard deviation, SD)], the number of assessable segments was 10.72±2.02 (SD). In the other 24 patients whose heart rates were above 70 bpm [mean 78.6 bpm±8.24 (SD)], the corresponding number was 9.75±1.74 (SD). No statistically significant difference was found in these two subgroups' t test, P>0.05. The new generation of 16-slice spiral CT with 0.37-s rotation time can satisfactorily evaluate the coronary arteries of patients with high heart rates (above 70 bpm, up to 102 bpm). (orig.)

  15. Oscillation measuring device for body of rotation

    International Nuclear Information System (INIS)

    Komita, Hideo.

    1994-01-01

    The present invention concerns an internal pump of a BWR type reactor and provides a device for detecting oscillations of a rotational shaft. Namely, recesses are formed along an identical circumference on the outer circumferential surface of the rotating portion each at a predetermined distance. The recesses rotate along with the rotation. An eddy current type displacement gage measures the distance to the outer circumferential surface of the rotating portion. The recesses are detected by the displacement gage as pulse signals. When the rotating portion oscillates, it is detected by the displacement gage as waveform signals. Accordingly, the output signals of the eddy current type displacement gage are formed by pulse signals superposed on the waveform signals. A rising detection circuit detects the rising position of the pulse signals as the components of the number of rotation of the rotating portion, and fall detection circuit detects the falling position. A comparator circuit is disposed in parallel with both of rising/falling detection circuits. A predetermined threshold value is set in the comparator circuit to output a signal when the inputted signal exceeds the value. (I.S.)

  16. A new irradiation unit constructed of self-moving gantry-CT and linac

    International Nuclear Information System (INIS)

    Kuriyama, Kengo; Onishi, Hiroshi; Sano, Naoki; Komiyama, Takafumi; Aikawa, Yoshihito; Tateda, Yoshihito; Araki, Tsutomu; Uematsu, Minoru

    2003-01-01

    Purpose: To improve reproducibility in stereotactic irradiation (STI) without using noninvasive immobilization devices or body frames, we have developed an integrated computed tomography (CT)-linac irradiation system connecting CT scanner and linac via a common treatment couch. Methods and Materials: This system consists of a linac, a CT scanner, and a common treatment couch. The linac and the CT gantry are positioned on opposite ends of the couch so that, by rotating the treatment couch, linac radiotherapy or CT scanning can be performed. The rotational axis of the linac gantry is coaxial with that of the CT gantry, and the position of the linac isocenter on the couch matches the origin of the coordinate system for CT scanning when the couch is rotated 180 degree sign toward the CT side. Instead of the couch moving into the gantry, as in conventional CT, in this case the table is fixed and scanning is accomplished by moving the gantry. We evaluated the rotational accuracy of the common couch and the scan-position accuracy of the self-moving gantry CT. Results: The positional accuracy of the common couch was 0.20, 0.18, and 0.39 mm in the lateral, longitudinal, and vertical directions, respectively. The scan-position accuracy of the CT gantry was less than 0.4 mm in the lateral, longitudinal, and vertical directions. Conclusion: This irradiation system has a high accuracy and is useful for noninvasive STI and for verification of the position of a target in three-dimensional conformal radiotherapy

  17. The Development and Clinical Use of a Beam ON-LINE PET System Mounted on a Rotating Gantry Port in Proton Therapy

    International Nuclear Information System (INIS)

    Nishio, Teiji; Miyatake, Aya; Ogino, Takashi; Nakagawa, Keiichi; Saijo, Nagahiro; Esumi, Hiroyasu

    2010-01-01

    Purpose: To verify the usefulness of our developed beam ON-LINE positron emission tomography (PET) system mounted on a rotating gantry port (BOLPs-RGp) for dose-volume delivery-guided proton therapy (DGPT). Methods and Materials: In the proton treatment room at our facility, a BOLPs-RGp was constructed so that a planar PET apparatus could be mounted with its field of view covering the iso-center of the beam irradiation system. Activity measurements were performed in 48 patients with tumors of the head and neck, liver, lungs, prostate, and brain. The position and intensity of the activity were measured using the BOLPs-RGp during the 200 s immediately after the proton irradiation. Results: The daily measured activity images acquired by the BOLPs-RGp showed the proton irradiation volume in each patient. Changes in the proton-irradiated volume were indicated by differences between a reference activity image (taken at the first treatment) and the daily activity-images. In the case of head-and-neck treatment, the activity distribution changed in the areas where partial tumor reduction was observed. In the case of liver treatment, it was observed that the washout effect in necrotic tumor cells was slower than in non-necrotic tumor cells. Conclusions: The BOLPs-RGp was developed for the DGPT. The accuracy of proton treatment was evaluated by measuring changes of daily measured activity. Information about the positron-emitting nuclei generated during proton irradiation can be used as a basis for ensuring the high accuracy of irradiation in proton treatment.

  18. 128-slice CT angiography of the aorta without ECG-gating: efficacy of faster gantry rotation time and iterative reconstruction in terms of image quality and radiation dose

    Energy Technology Data Exchange (ETDEWEB)

    Russo, Vincenzo; Garattoni, Monica; Buia, Francesco; Attina, Domenico; Lovato, Luigi; Zompatori, Maurizio [University Hospital ' ' S.Orsola' ' , Cardio-Thoracic-Vascular Department, Cardio-Thoracic Radiology Unit, Bologna (Italy)

    2016-02-15

    To evaluate image quality and radiation dose of non ECG-gated 128-slice CT angiography of the aorta (CTAA) with fast gantry rotation time and iterative reconstruction. Four hundred and eighty patients underwent non ECG-gated CTAA. Qualitative and quantitative image quality assessments were performed. Radiation dose was assessed and compared with the dose of patients who underwent ECG-gated CTAA (n = 126) and the dose of previous CTAA performed with another CT (n = 339). Image quality (aortic root-ascending portion) was average-to-excellent in more than 94 % of cases, without any non-diagnostic scan. For proximal coronaries, image quality was average-to-excellent in more than 50 %, with only 21.5 % of non-diagnostic cases. Quantitative analysis results were also good. Mean radiation dose for thoracic CTAA was 5.6 mSv versus 20.6 mSv of ECG-gated protocol and 20.6 mSv of 16-slice CTAA scans, with an average dose reduction of 72.8 % (p < 0.001). Mean radiation dose for thoracic-abdominal CTAA was 9.7 mSv, versus 20.9 mSv of 16-slice CTAA scans, with an average dose reduction of 53.6 % (p < 0.001). Non ECG-gated 128-slice CTAA is feasible and able to provide high quality visualization of the entire aorta without significant motion artefacts, together with a considerable dose and contrast media volume reduction. (orig.)

  19. 128-slice CT angiography of the aorta without ECG-gating: efficacy of faster gantry rotation time and iterative reconstruction in terms of image quality and radiation dose

    International Nuclear Information System (INIS)

    Russo, Vincenzo; Garattoni, Monica; Buia, Francesco; Attina, Domenico; Lovato, Luigi; Zompatori, Maurizio

    2016-01-01

    To evaluate image quality and radiation dose of non ECG-gated 128-slice CT angiography of the aorta (CTAA) with fast gantry rotation time and iterative reconstruction. Four hundred and eighty patients underwent non ECG-gated CTAA. Qualitative and quantitative image quality assessments were performed. Radiation dose was assessed and compared with the dose of patients who underwent ECG-gated CTAA (n = 126) and the dose of previous CTAA performed with another CT (n = 339). Image quality (aortic root-ascending portion) was average-to-excellent in more than 94 % of cases, without any non-diagnostic scan. For proximal coronaries, image quality was average-to-excellent in more than 50 %, with only 21.5 % of non-diagnostic cases. Quantitative analysis results were also good. Mean radiation dose for thoracic CTAA was 5.6 mSv versus 20.6 mSv of ECG-gated protocol and 20.6 mSv of 16-slice CTAA scans, with an average dose reduction of 72.8 % (p < 0.001). Mean radiation dose for thoracic-abdominal CTAA was 9.7 mSv, versus 20.9 mSv of 16-slice CTAA scans, with an average dose reduction of 53.6 % (p < 0.001). Non ECG-gated 128-slice CTAA is feasible and able to provide high quality visualization of the entire aorta without significant motion artefacts, together with a considerable dose and contrast media volume reduction. (orig.)

  20. Preliminary design study for a corkscrew gantry

    International Nuclear Information System (INIS)

    Koehler, A.M.

    1987-01-01

    For two years or more a group including the author has been working together to study some problems related to the design of a gentry system for flexible direction of a proton beam for clinical treatments. Some consideration was given to the classic gantry geometry. Attempting to reduce the radius of the gantry arm by reducing the drift space after the scattering foils led to an analysis of the significance of inverse square intensity effects. The conclusion reached is that a drift space of about 3 meters is required to preserve some skin sparing for larger targets. To circumvent this problem the scattering foils ere put somewhere inside or even before the gantry system, accepting the fact that magnet apertures would have to be increased. This gantry system has the interesting ability to produce oblong fields of excellent uniformity with reasonable efficiency, preferentially with the long axis of the field parallel to the axis of rotation. It was disappointing, however, to find that the overall size of the gantry with its counterweights remained very large. Another change in geometry was proposed therefore in order to reduce the space taken up by the gantry and its counterweight. The beam is bent 45 0 in the horizontal plane and then again by 45 0 so that it is pointing away from isocenter, but in the plan of rotation of the gantry. The beam is now bent in that plane of rotation until it is pointed at isocenter. This is accomplished by two bends of 135 0 each with a suitable drift space between them so that the beam is pointed vertically downward at isocenter. The three dimensional complexity of the beam trajectory led to the name Corkscrew Gantry

  1. Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle

    International Nuclear Information System (INIS)

    Clarke, M F; Budgell, G J

    2008-01-01

    Amorphous silicon electronic portal imaging devices (EPIDs) are used to perform routine quality control (QC) checks on the multileaf collimators (MLCs) at this centre. Presently, these checks are performed at gantry angle 0 0 and are considered to be valid for all other angles. Since therapeutic procedures regularly require the delivery of MLC-defined fields to the patient at a wide range of gantry angles, the accuracy of the QC checks at other gantry angles has been investigated. When the gantry is rotated to angles other than 0 0 it was found that the apparent pixel size measured using the EPID varies up to a maximum value of 0.0015 mm per pixel due to a sag in the EPID of up to 9.2 mm. A correction factor was determined using two independent methods at a range of gantry angles between 0 deg. and 360 deg. The EPID was used to measure field sizes (defined by both x-jaws and MLC) at a range of gantry angles and, after this correction had been applied, any residual gravitational sag was studied. It was found that, when fields are defined by the x-jaws and y-back-up jaws, no errors of greater than 0.5 mm were measured and that these errors were no worse when the MLC was used. It was therefore concluded that, provided the correction is applied, measurements of the field size are, in practical terms, unaffected by gantry angle. Experiments were also performed to study how the reproducibility of individual leaves is affected by gantry angle. Measurements of the relative position of each individual leaf (minor offsets) were performed at a range of gantry angles and repeated three times. The position reproducibility was defined by the RMS error in the position of each leaf and this was found to be 0.24 mm and 0.21 mm for the two leaf banks at a gantry angle of 0 0 . When measurements were performed at a range of gantry angles, these reproducibility values remained within 0.09 mm and 0.11 mm. It was therefore concluded that the calibration of the Elekta MLC is stable at

  2. 'Riesenrad' ion gantry for hadrontherapy: Part III

    International Nuclear Information System (INIS)

    Benedikt, M.; Bryant, P.; Holy, P.; Pullia, M.

    1999-01-01

    When using accelerator beams for cancer therapy, the three-dimensional freedom afforded by a gantry helps the treatment planner to spread out surface doses, avoid directions that intercept vital organs and irradiate a volume that is conformal with the tumour. The general preference is for an iso-centric gantry turning 360 deg. in the vertical plane around the patient bed with sufficient space to be able to orientate the patient through 360 deg. in the horizontal plane. For hadrontherapy, gantries are impressive structures of the order of 10 m in diameter and 100 ton in weight and to date only proton gantries have been demonstrated to operate satisfactorily. The increased magnetic rigidity of say carbon ions will make ion gantries more difficult and costly to build. For this reason, exo-centric gantries and, in particular the so-called 'Riesenrad' gantry with a single 90 deg. bending magnet, merit further attention. The power consumption is reduced and the heavy magnets with their counterbalance weight are reduced and are kept close to the axis. The treatment room, which is lighter, is positioned at a larger radius, but only the patient bed requires careful alignment. An optics module called a 'rotator' is needed to match an incoming dispersion vector to the gantry in order to have an achromatic beam at the patient. A practical design is described that assumes the beam is derived from a slow-extraction scheme in a synchrotron and that the beam sizes are controlled by modules in the transfer line. Magnetic scanning is integrated into the gantry optics for both transverse directions

  3. `Riesenrad' ion gantry for hadrontherapy: Part III

    Science.gov (United States)

    Benedikt, M.; Bryant, P.; Holy, P.; Pullia, M.

    1999-07-01

    When using accelerator beams for cancer therapy, the three-dimensional freedom afforded by a gantry helps the treatment planner to spread out surface doses, avoid directions that intercept vital organs and irradiate a volume that is conformal with the tumour. The general preference is for an iso-centric gantry turning 360° in the vertical plane around the patient bed with sufficient space to be able to orientate the patient through 360° in the horizontal plane. For hadrontherapy, gantries are impressive structures of the order of 10 m in diameter and 100 ton in weight and to date only proton gantries have been demonstrated to operate satisfactorily. The increased magnetic rigidity of say carbon ions will make ion gantries more difficult and costly to build. For this reason, exo-centric gantries and, in particular the so-called `Riesenrad' gantry with a single 90° bending magnet, merit further attention. The power consumption is reduced and the heavy magnets with their counterbalance weight are reduced and are kept close to the axis. The treatment room, which is lighter, is positioned at a larger radius, but only the patient bed requires careful alignment. An optics module called a `rotator' is needed to match an incoming dispersion vector to the gantry in order to have an achromatic beam at the patient. A practical design is described that assumes the beam is derived from a slow-extraction scheme in a synchrotron and that the beam sizes are controlled by modules in the transfer line. Magnetic scanning is integrated into the gantry optics for both transverse directions.

  4. Detection and correction for EPID and gantry sag during arc delivery using cine EPID imaging.

    Science.gov (United States)

    Rowshanfarzad, Pejman; Sabet, Mahsheed; O'Connor, Daryl J; McCowan, Peter M; McCurdy, Boyd M C; Greer, Peter B

    2012-02-01

    Electronic portal imaging devices (EPIDs) have been studied and used for pretreatment and in-vivo dosimetry applications for many years. The application of EPIDs for dosimetry in arc treatments requires accurate characterization of the mechanical sag of the EPID and gantry during rotation. Several studies have investigated the effects of gravity on the sag of these systems but each have limitations. In this study, an easy experiment setup and accurate algorithm have been introduced to characterize and correct for the effect of EPID and gantry sag during arc delivery. Three metallic ball bearings were used as markers in the beam: two of them fixed to the gantry head and the third positioned at the isocenter. EPID images were acquired during a 360° gantry rotation in cine imaging mode. The markers were tracked in EPID images and a robust in-house developed MATLAB code was used to analyse the images and find the EPID sag in three directions as well as the EPID + gantry sag by comparison to the reference gantry zero image. The algorithm results were then tested against independent methods. The method was applied to compare the effect in clockwise and counter clockwise gantry rotations and different source-to-detector distances (SDDs). The results were monitored for one linear accelerator over a course of 15 months and six other linear-accelerators from two treatment centers were also investigated using this method. The generalized shift patterns were derived from the data and used in an image registration algorithm to correct for the effect of the mechanical sag in the system. The Gamma evaluation (3%, 3 mm) technique was used to investigate the improvement in alignment of cine EPID images of a fixed field, by comparing both individual images and the sum of images in a series with the reference gantry zero image. The mechanical sag during gantry rotation was dependent on the gantry angle and was larger in the in-plane direction, although the patterns were not

  5. SU-E-T-64: CG-Based Radiation Therapy Simulator with Physical Modeling for Avoidance of Collisions Between Gantry and Couch Or Patient

    International Nuclear Information System (INIS)

    Yamanouchi, M; Arimura, H; Yuda, I

    2014-01-01

    Purpose: It is time-consuming and might cause re-planning to check couch-gantry and patient-gantry collisions on a radiotherapy machine when using couch rotations for non-coplanar beam angles. The aim of this study was to develop a computer-graphics (CG)-based radiation therapy simulator with physical modeling for avoidance of collisions between gantry and couch or patient on a radiotherapy machine. Methods: The radiation therapy simulator was three-dimensionally constructed including a radiotherapy machine (Clinac iX, Varian Medical Systems), couch, and radiation treatment room according to their designs by using a physical-modeling-based computer graphics software (Blender, free and open-source). Each patient was modeled by applying a surface rendering technique to their planning computed tomography (CT) images acquired from 16-slice CT scanner (BrightSpeed, GE Healthcare). Immobilization devices for patients were scanned by the CT equipment, and were rendered as the patient planning CT images. The errors in the collision angle of the gantry with the couch or patient between gold standards and the estimated values were obtained by fixing the gantry angle for the evaluation of the proposed simulator. Results: The average error of estimated collision angles to the couch head side was -8.5% for gantry angles of 60 to 135 degree, and -5.5% for gantry angles of 225 to 300 degree. Moreover, the average error of estimated collision angles to the couch foot side was -1.1% for gantry angles of 60 to 135 degree, and 1.4% for gantry angles of 225 to 300 degree. Conclusion: The CG-based radiation therapy simulator could make it possible to estimate the collision angle between gantry and couch or patient on the radiotherapy machine without verifying the collision angles in the radiation treatment room

  6. Design of an achromatic and uncoupled medical gantry for radiation therapy

    International Nuclear Information System (INIS)

    Tsoupas, N.; Kayran, D.; Litvinenko, V.; MacKay, W.W.

    2011-01-01

    We are presenting the layout and the optics of a beam line to be used as a medical gantry in radiation therapy. The optical properties of the gantry's beam line are such as to make the beam line achromatic and uncoupled. These two properties make the beam spot size, which is delivered and focused by the gantry, on the tumor of the patient, independent of the angular orientation of the gantry. In this paper we present the layout of the magnetic elements of the gantry, and also present the theoretical basis for the optics design of such a gantry. A medical gantry, as it is used in the radiation treatment of cancer patients, is the last part of the beam optical system, of the accelerator complex, which delivers and focuses the beam on the tumor. The curved line shown in figure 1 is a schematic diagram of a gantry which can rotate about a horizontal axis. The particle beam (green arrow in fig. 1) enters the gantry, and is guided by the gantry on the tumor (red spot in fig. 1). As the gantry rotates about the axis shown in figure 1, the beam exiting the gantry always lies on a plane normal to the rotation axis at the point of the icocenter. Thus the gantry facilitates the ability of the beam delivery system, to deliver the beam at the tumor, which is placed at the icocenter, from any angle on this vertical plane, which is normal to the rotation angle of the gantry as stated earlier. The gantry consists of dipoles and quadrupoles elements whose median symmetry plane lies on a plane which contains the rotation axis of the gantry. In this paper we define this plane as the 'plane of the gantry'. As the beam is transported along the axis of rotation of the gantry and before it enters the gantry, it is focused by 'normal' quadrupoles and experiences no linear beam coupling. Subsequently the beam enters the gantry, and is transported by the gantry to the delivery point which is the tumor. The transported beam at the tumor is still linearly uncoupled as long as the plane of the

  7. Preliminary Test on Hydraulic Rotation Device for Neutron Transmutation Doping

    International Nuclear Information System (INIS)

    Park, Ki-Jung; Kang, Han-Ok; Kim, Seong Hoon; Park, Cheol

    2014-01-01

    The Korea Atomic Energy Research Institute (KAERI) is developing a new Research Reactor (KJRR) which will be located at KIJANG in the south-eastern province of Korea. The KJRR will be mainly utilized for isotope production, NTD production, and the related research activities. During the NTD process, the irradiation rig containing the silicon ingot rotates at the constant speed to ensure precisely defined homogeneity of the irradiation. A new NTD Hydraulic Rotation Device (NTDHRD) is being developed to rotate the irradiation rigs at the required speed. In this study, the preliminary test and the analysis for the rotation characteristic of the NTDHRD, which is developed through the conceptual design, are described. A new NTD hydraulic rotation device is being developed for the purpose of application to the KIJANG research reactor (KJRR). The preliminary test and analysis for the rotation characteristic of the NTDHRD, which is developed through the conceptual design, are conducted in experimental apparatus. The film thickness by the thrust bearing is measured and the minimum required mass flow rate for stable rotation is determined

  8. Volumetric modulated arc therapy: IMRT in a single gantry arc

    International Nuclear Information System (INIS)

    Otto, Karl

    2008-01-01

    In this work a novel plan optimization platform is presented where treatment is delivered efficiently and accurately in a single dynamically modulated arc. Improvements in patient care achieved through image-guided positioning and plan adaptation have resulted in an increase in overall treatment times. Intensity-modulated radiation therapy (IMRT) has also increased treatment time by requiring a larger number of beam directions, increased monitor units (MU), and, in the case of tomotherapy, a slice-by-slice delivery. In order to maintain a similar level of patient throughput it will be necessary to increase the efficiency of treatment delivery. The solution proposed here is a novel aperture-based algorithm for treatment plan optimization where dose is delivered during a single gantry arc of up to 360 deg. The technique is similar to tomotherapy in that a full 360 deg. of beam directions are available for optimization but is fundamentally different in that the entire dose volume is delivered in a single source rotation. The new technique is referred to as volumetric modulated arc therapy (VMAT). Multileaf collimator (MLC) leaf motion and number of MU per degree of gantry rotation is restricted during the optimization so that gantry rotation speed, leaf translation speed, and dose rate maxima do not excessively limit the delivery efficiency. During planning, investigators model continuous gantry motion by a coarse sampling of static gantry positions and fluence maps or MLC aperture shapes. The technique presented here is unique in that gantry and MLC position sampling is progressively increased throughout the optimization. Using the full gantry range will theoretically provide increased flexibility in generating highly conformal treatment plans. In practice, the additional flexibility is somewhat negated by the additional constraints placed on the amount of MLC leaf motion between gantry samples. A series of studies are performed that characterize the relationship

  9. Design of a "Riesenrad" Ion Gantry for Hadrontherapy

    CERN Document Server

    Reimoser, S

    2000-01-01

    The benefit of hadrontherapy can be maximized by offering the possibility to deliver the particle beam from any direction in space towards the patient with the help of a medical gantry. For carbon ions, their increased (magnetic) beam rigidity yields considerable structural difficulties and has so far prevented a practical realization of an ion gantry. The concept of a 'Riesenrad' ion gantry promises to provide an effective and efficient solution. The basic idea is to deflect the ion beam with a single 90° dipole, which rotates around the incoming beam axis, and direct it towards the eccentrically positioned patient cabin. Inside the cabin similar conditions as exist in a classical isocentric treatment room prevail. The practical design of such a Riesenrad gantry, its structural principles and its function are presented. The underlying beam optics and its integration into the mechanical structure are explained. Aspects of safety, and flexibility are discussed.

  10. The PSI Gantry 2: a second generation proton scanning gantry.

    Science.gov (United States)

    Pedroni, Eros; Bearpark, Ralph; Böhringer, Terence; Coray, Adolf; Duppich, Jürgen; Forss, Sven; George, David; Grossmann, Martin; Goitein, Gudrun; Hilbes, Christian; Jermann, Martin; Lin, Shixiong; Lomax, Antony; Negrazus, Marco; Schippers, Marco; Kotle, Goran

    2004-01-01

    PSI is still the only location in which proton therapy is applied using a dynamic beam scanning technique on a very compact gantry. Recently, this system is also being used for the application of intensity-modulated proton therapy (IMPT). This novel technical development and the success of the proton therapy project altogether have led PSI in Year 2000 to further expand the activities in this field by launching the project PROSCAN. The first step is the installation of a dedicated commercial superconducting cyclotron of a novel type. The second step is the development of a new gantry, Gantry 2. For Gantry 2 we have chosen an iso-centric compact gantry layout. The diameter of the gantry is limited to 7.5 m, less than in other gantry systems (approximately 10-12 m). The space in the treatment room is comfortably large, and the access on a fixed floor is possible any time around the patient table. Through the availability of a faster scanning system, it will be possible to treat the target volume repeatedly in the same session. For this purpose, the dynamic control of the beam intensity at the ion source and the dynamic variation of the beam energy will be used directly for the shaping of the dose.

  11. Development of signal acquisition device of rotating coil measurement system

    International Nuclear Information System (INIS)

    Zhou Jianxin; Li Li; Kang Wen; Deng Chengdong; Yin Baogui; Fu Shinian

    2013-01-01

    A new rotating coil magnetic measurement system using the technical solution of the combination of a dynamic signal acquisition card and software with specific functions was developed. The acquisition device of the system successfully implemented the function of the PDI-5025 integrator. The sampling rate, the range, the accuracy and the flexibility of the system were improved. The development program of signal acquisition equipment, the realization of the acquisition function and the reliability and stability of the system were introduced. (authors)

  12. {100} or 45.degree.-rotated {100}, semiconductor-based, large-area, flexible, electronic devices

    Science.gov (United States)

    Goyal, Amit [Knoxville, TN

    2012-05-15

    Novel articles and methods to fabricate the same resulting in flexible, {100} or 45.degree.-rotated {100} oriented, semiconductor-based, electronic devices are disclosed. Potential applications of resulting articles are in areas of photovoltaic devices, flat-panel displays, thermophotovoltaic devices, ferroelectric devices, light emitting diode devices, computer hard disc drive devices, magnetoresistance based devices, photoluminescence based devices, non-volatile memory devices, dielectric devices, thermoelectric devices and quantum dot laser devices.

  13. Development and engineering design of a novel exocentric carbon-ion gantry for cancer therapy (the 'Riesenrad' gantry)

    International Nuclear Information System (INIS)

    Reimoser, S.A.

    2000-10-01

    The objective of the present thesis is to develop a novel gantry concept leading to a viable and efficient design for a carbon-ion gantry. Chapter 1 is meant to give some background information on radiation therapy and on the promises of ion treatment. Chapter 2 presents and summarizes existing gantry systems (in particular for protons) and briefly describes the experimental facilities for ion treatment in order to form an idea in what direction the design of an ion gantry might point. Chapter 3 is dedicated to the definition of the medical, beam-optical and mechanical constraints affecting the design of the ion gantry. The principal processes likely to take place inside the ion gantry are highlighted and the various technical systems and their interdependencies are investigated. As a result of these considerations it was decided to focus on the development of an exocentric gantry system for ion therapy, which was eventually named the 'Riesenrad' gantry. In contrast to conventional isocentric gantries, the main 90 o -bending magnet of the Riesenrad is placed on the axis of gantry rotation, hence minimizing the moment of inertia of the mobile structure and maximizing its rigidity. Several preliminary structural concepts for the Riesenrad gantry were generated, analyzed and compared in chapter 4. Finally, the variant featuring an independent telescopic cabin was chosen for further elaboration in a design phase (chapter 5). In the proposed variant, the bending magnet that sits on the axis and its counterweight (62 ton and 23 ton respectively) are supported by a central 'cage' of about 40 t. This gantry yields a higher efficiency in terms of structural weight to supported load compared to any other possible variant. A patient cabin is smoothly moved towards the desired treatment position by a system that is mechanically de-coupled from the central cage. Any desired position (that corresponds to a particular gantry angle) is achieved by vertical translation and

  14. Particle Trajectories in Rotating Wall Cell Culture Devices

    Science.gov (United States)

    Ramachandran N.; Downey, J. P.

    1999-01-01

    Cell cultures are extremely important to the medical community since such cultures provide an opportunity to perform research on human tissue without the concerns inherent in experiments on individual humans. Development of cells in cultures has been found to be greatly influenced by the conditions of the culture. Much work has focused on the effect of the motions of cells in the culture relative to the solution. Recently rotating wall vessels have been used with success in achieving improved cellular cultures. Speculation and limited research have focused on the low shear environment and the ability of rotating vessels to keep cells suspended in solution rather than floating or sedimenting as the primary reasons for the improved cellular cultures using these devices. It is widely believed that the cultures obtained using a rotating wall vessel simulates to some degree the effect of microgravity on cultures. It has also been speculated that the microgravity environment may provide the ideal acceleration environment for culturing of cellular tissues due to the nearly negligible levels of sedimentation and shear possible. This work predicts particle trajectories of cells in rotating wall vessels of cylindrical and annular design consistent with the estimated properties of typical cellular cultures. Estimates of the shear encountered by cells in solution and the interactions with walls are studied. Comparisons of potential experiments in ground and microgravity environments are performed.

  15. Design, fabrication and transportation of Si rotating device

    International Nuclear Information System (INIS)

    Kimura, Nobuaki; Imaizumi, Tomomi; Takemoto, Noriyuki; Tanimoto, Masataka; Saito, Takashi; Hori, Naohiko; Tsuchiya, Kunihiko; Romanova, Nataliya; Gizatulin, Shamil; Martyushov, Alexandr; Nakipov, Darkhan; Chakrov, Petr; Tanaka, Futoshi; Nakajima, Takeshi

    2012-06-01

    Si semiconductor production by Neutron Transmutation Doping (NTD) method using the Japan Materials Testing Reactor (JMTR) has been investigated in Neutron Irradiation and Testing Reactor Center, Japan Atomic Energy Agency (JAEA) in order to expand industry use. As a part of investigations, irradiation test of silicon ingot for development of NTD-Si with high quality was planned using WWR-K in Institute of Nuclear Physics (INP), National Nuclear Center of Republic of Kazakhstan (NNC-RK) based on one of specific topics of cooperation (STC), Irradiation Technology for NTD-Si (STC No.II-4), on the implementing arrangement between NNC-RK and the JAEA for 'Nuclear Technology on Testing/Research Reactors' in cooperation in research and development in nuclear energy and technology. As for the irradiation test, Si rotating device was fabricated in JAEA, and the fabricated device was transported with irradiation specimens from JAEA to INP-NNC-RK. This report described the design, the fabrication, the performance test of the Si rotating device and transportation procedures. (author)

  16. SU-E-T-160: Characterization and Monitoring of Linear Accelerator Gantry Radiation Isocenter Motion

    International Nuclear Information System (INIS)

    Letourneau, D; Amin, N; Wang, K; Norrlinger, B; Jaffray, D; McNiven, A

    2015-01-01

    Purpose: To characterize the motion of the radiation isocenter, over time, as a function of gantry rotation for multiple linear accelerators (linacs). Two semi-automated image-based quality control (QC) test workflows were designed to achieve this goal. Methods: The full QC-test workflow for motion characterization consisted of acquiring 16 megavoltage images at 8 gantry angles of a ball-bearing suspended off the end of the couch. Performance constancy was assessed using a shortened QC-test workflow which consisted of imaging a cube phantom placed on the couch (5 images at 4 gantry angles). Both workflows use an image processing algorithm to determine the field center and phantom position on each image and computed radiation isocenter motion as a function of gantry angle. Motion was characterized for 9 linacs of same model and performance monitored for 2 months on 3 linacs. Results: The maximum isocenter motion determined with the full-workflow for 9 linacs was within 0.38–0.79 mm. The shortened-workflow usually agreed within 0.1 mm with the full-workflow and the time required for these methods was about 4 and 15 min, respectively. For all linacs, the isocenter motion perpendicular to the gantry rotation plane followed a consistent pattern with maximum amplitude of 0.36–0.59 mm. In the gantry rotation plane, the variation among linacs was higher and the beam axis described a circle of up to 0.6 mm radius around the gantry axis of rotation (2 linacs). The radiation isocenter motion was stable as a function of time for the monitored linacs and was within ±0.1 mm of the average. Conclusion: Radiation isocenter motion parallel and perpendicular to the gantry rotation plane was characterized. In the gantry rotation plane, beam spot positioning adjustment might be used to reduce the observed radiation isocenter motion. A shortened-workflow was designed and enables performance monitoring over time

  17. Motion planning for gantry mounted manipulators

    DEFF Research Database (Denmark)

    Olsen, Anders Lau; Petersen, Henrik Gordon

    2007-01-01

    We present a roadmap based planner for finding robot motions for gantry mounted manipulators for a line welding application at Odense Steel Shipyard (OSS). The robot motions are planned subject to constraints on when the gantry may be moved. We show that random sampling of gantry configurations...

  18. Visualization of coronary arteries in CT as assessed by a new 16 slice technology and reduced gantry rotation time: first experiences; Darstellung der Herzkranzgefaesse im CT mittels neuer 16-Zeilen-Technologie und reduzierter Rotationszeit: erste Erfahrungen

    Energy Technology Data Exchange (ETDEWEB)

    Heuschmid, M.; Kuettner, A.; Kopp, A.F.; Claussen, C.D. [Abt. Radiologische Diagnostik, Universitaetsklinikum Tuebingen (Germany); Flohr, T.; Schaller, S.; Hartung, A.; Ohnesorge, B. [Siemens Medical Solutions, Forchheim (Germany); Wildberger, J.E. [Klinik fuer Radiologische Diagnostik, Universitaetsklinikum (RWTH) Aachen (Germany); Lell, M.; Baum, U. [Abt. Innere Medizin IV (Kardiologie), Universitaetsklinikum Tuebingen (Germany); Schroeder, S. [Inst. fuer Diagnostische Radiologie, Universitaetsklinikum Erlangen (Germany)

    2002-06-01

    Purpose: First evaluation of image quality of a new 16-slice multidetector-row computed tomography (MDCT) for the assessment of coronary artery disease and lesion detection of the coronary arteries. Materials and Methods: On a newly developed 16-slice CT scanner (SOMATOM Sensation 16, Siemens, Forchheim, Germany) a calcium score as well as a contrast-enhanced CT angiography (CTA) were performed on 4 patients with retrospective ECG-gating and a gantry rotation time of 420 ms to exclude or follow-up coronary heart disease. CTA was performed after injecting 120 ml contrast media intravenously. After medication with a {beta}-Blocker, the heart rate was between 55 and 67 bpm. Results: The scan time for calcium score was 12s, for CTA 18s (scan range 15 and 12 cm, respectively). Volume score was between 0 and 256.4. In the CT angiography the entire coronary tree could be visualized in all patients up to the very distal subsegmental branches. In two patients a complete occlusion of the RCA and the LAD were depicted, respectively. In one of these patients, a large aneurysm of the left anterior ventricular wall was also delineated. Conclusion: Considering our first experiences with the new 16-slice technology, an excellent visualization of the entire coronary tree including the very distal and side branches due to substantially increased spatial resolution seems to be achievable. In these patients the acquired image quality raises the hope for improved, non-invasive cardiac diagnostics. In larger studies, the clinical impact of this new technology needs to be further investigated. (orig.) [German] Ziel: Darstellung erster Ergebnisse zur Bildqualitaet eines neuen 16-Zeilen Multidetektor-Computertomographen (MDCT) bei der Diagnostik der koronaren Herzerkrankung (KHK) und hochgradiger Stenosen der Herzkranzgefaesse. Material und Methode: Mit einem neuen 16-Zeilen-Computertomographen (SOMATOM Sensation 16, Siemens, Forchheim) wurde bei 4 Patienten zum Ausschluss oder

  19. "Riesenrad" Ion Gantry for Hadron Therapy, 3

    OpenAIRE

    Benedikt, Michael; Bryant, P J; Holy, P; Pullia, M

    1999-01-01

    When using accelerator beams for cancer therapy, the three-dimensional freedom afforded by a gantry helps the treatment planner to spread out surface doses, avoid directions that intercept vital organs and irradiate a volume that is conformal with the tumour. The general preference is for an iso-centric gantry turning 360° in the vertical plane around the patient bed with sufficient space to be able to orientate the patient through 360° in the horizontal plane. For hadrontherapy, gantries are...

  20. A feasibility study of Dynamic Phantom scanner for quality assurance of photon beam profiles at various gantry angles.

    Science.gov (United States)

    Zhang, Yunkai; Hsi, Wen C; Chu, James C H; Bernard, Damian B; Abrams, Ross A

    2005-01-01

    ) were observed. Additional profiles with the gantry at 90 degrees and 270 degrees were performed for both MLC- and jaw-shaped photon beams and electron beams to evaluate the effect of gantry rotation. General good agreement is seen (less than 1 % variation) at all field sizes for collimator-shaped 6 MV and 18 MV photon beams. Similar variations observed for MLC-shaped photon beams indicate that the uncertainty in MLC position is similar to that for the collimator jaws. We conclude that the Dynamic Phantom scanner is a useful device for the routine quality assurance on beam profiles of photon beams and for constancy check on electron beams at various gantry angles. Caution should be taken when using this device to acquire basic electron dosimetry data.

  1. A feasibility study of the Dynamic Phantom scanner for quality assurance of beam profiles at various gantry angles

    Science.gov (United States)

    Zhang, Yunkai; Hsi, Wen C.; Chu, James C.H.; Bernard, Damian B.; Abrams, Ross A.

    2005-01-01

    ) were observed. Additional profiles with the gantry at 90° and 270° were performed for both MLC‐ and jaw‐shaped photon beams and electron beams to evaluate the effect of gantry rotation. General good agreement is seen (less than 1 % variation) at all field sizes for collimator‐shaped 6 MV and 18 MV photon beams. Similar variations observed for MLC‐shaped photon beams indicate that the uncertainty in MLC position is similar to that for the collimator jaws. We conclude that the Dynamic Phantom scanner is a useful device for the routine quality assurance on beam profiles of photon beams and for constancy check on electron beams at various gantry angles. Caution should be taken when using this device to acquire basic electron dosimetry data. PACS number: 87.53.‐j PMID:15940212

  2. "Riesenrad" Ion Gantry for Hadron Therapy, 3

    CERN Document Server

    Benedikt, Michael; Holy, P; Pullia, M

    1999-01-01

    When using accelerator beams for cancer therapy, the three-dimensional freedom afforded by a gantry helps the treatment planner to spread out surface doses, avoid directions that intercept vital organs and irradiate a volume that is conformal with the tumour. The general preference is for an iso-centric gantry turning 360° in the vertical plane around the patient bed with sufficient space to be able to orientate the patient through 360° in the horizontal plane. For hadrontherapy, gantries are impressive structures of the order of 10 m in diameter and 100 tons in weight and to date only proton gantries have been demonstrated to operate satisfactorily. The increased magnetic rigidity of say carbon ions will make ion gantries more difficult and costly to build. For this reason, exo-centric gantries and, in particular the so-called 'Riesenrad' gantry with a single 90° bending magnet, merit further attention. The power consumption is reduced and the heavy magnets with their counterbalance weight are reduced and...

  3. Emplacement Gantry Gap Analysis Study

    International Nuclear Information System (INIS)

    Thornley, R.

    2005-01-01

    To date, the project has established important to safety (ITS) performance requirements for structures, systems, and components (SSCs) based on the identification and categorization of event sequences that may result in a radiological release. These performance requirements are defined within the ''Nuclear Safety Design Bases for License Application'' (NSDB) (BSC 2005 [DIRS 171512], Table A-11). Further, SSCs credited with performing safety functions are classified as ITS. In turn, assurance that these SSCs will perform as required is sought through the use of consensus codes and standards. This gap analysis is based on the design completed for license application only. Accordingly, identification of ITS SSCs beyond those defined within the NSDB are based on designs that may be subject to further development during detail design. Furthermore, several design alternatives may still be under consideration to satisfy certain safety functions, and final selection will not be determined until further design development has occurred. Therefore, for completeness, alternative designs currently under consideration will be discussed throughout this study. This gap analysis will evaluate each code and standard identified within the ''Emplacement Gantry ITS Standards Identification Study'' (BSC 2005 [DIRS 173586]) to ensure each ITS performance requirement is fully satisfied. When a performance requirement is not fully satisfied, a gap is highlighted. This study will identify requirements to supplement or augment the code or standard to meet performance requirements. Further, this gap analysis will identify nonstandard areas of the design that will be subject to a design development plan. Nonstandard components and nonstandard design configurations are defined as areas of the design that do not follow standard industry practices or codes and standards. Whereby, assurance that an SSC will perform as required may not be readily sought though the use of consensus standards. This

  4. Shaf rotation speed measurement device, its checking and variations checking

    International Nuclear Information System (INIS)

    Gadrault, Robert.

    1976-01-01

    Appliance for measuring the rotational speed of a shaft and monitoring of this speed and its changes. The uses to be made specifically concern the nuclear field and in this field the drive shafts of water coolant feed pumps. Detecting te rotation of the shaft concerned may be resolved with electronic sensors or proximity detectors which, because they are not in mechanical connexion with the shaft the speed of which they are to help detect, do not bring any lack of precision. The accuracy of the determination them depends only on the downstream processing of the data supplied by the sensor [fr

  5. An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments

    International Nuclear Information System (INIS)

    Rowshanfarzad, Pejman; McGarry, Conor K; Barnes, Michael P; Sabet, Mahsheed; Ebert, Martin A

    2014-01-01

    In modern radiotherapy, it is crucial to monitor the performance of all linac components including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method has been introduced in conjunction with an algorithm to investigate the stability of these systems during arc treatments with the aim of ensuring the accuracy of linac mechanical performance. The Varian EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in MLC carriages as a result of linac rotation were separately investigated by acquisition of EPID images of a simple phantom comprised of 5 ball-bearings during arc delivery. A fast and robust software package was developed for automated analysis of image data. Twelve Varian linacs of different models were investigated. The average EPID sag was within 1 mm for all tested linacs. All machines showed less than 1 mm gantry sag. Changes in SDD values were within 1.7 mm except for three linacs of one centre which were within 9 mm. Values of EPID skewness and tilt were negligible in all tested linacs. The maximum sag in MLC leaf bank assemblies was around 1 mm. The EPID sag showed a considerable improvement in TrueBeam linacs. The methodology and software developed in this study provide a simple tool for effective investigation of the behaviour of linac components with gantry rotation. It is reproducible and accurate and can be easily performed as a routine test in clinics

  6. Novel AC Servo Rotating and Linear Composite Driving Device for Plastic Forming Equipment

    Science.gov (United States)

    Liang, Jin-Tao; Zhao, Sheng-Dun; Li, Yong-Yi; Zhu, Mu-Zhi

    2017-07-01

    The existing plastic forming equipment are mostly driven by traditional AC motors with long transmission chains, low efficiency, large size, low precision and poor dynamic response are the common disadvantages. In order to realize high performance forming processes, the driving device should be improved, especially for complicated processing motions. Based on electric servo direct drive technology, a novel AC servo rotating and linear composite driving device is proposed, which features implementing both spindle rotation and feed motion without transmission, so that compact structure and precise control can be achieved. Flux switching topology is employed in the rotating drive component for strong robustness, and fractional slot is employed in the linear direct drive component for large force capability. Then the mechanical structure for compositing rotation and linear motion is designed. A device prototype is manufactured, machining of each component and the whole assembly are presented respectively. Commercial servo amplifiers are utilized to construct the control system of the proposed device. To validate the effectiveness of the proposed composite driving device, experimental study on the dynamic test benches are conducted. The results indicate that the output torque can attain to 420 N·m and the dynamic tracking errors are less than about 0.3 rad in the rotating drive. the dynamic tracking errors are less than about 1.6 mm in the linear feed. The proposed research provides a method to construct high efficiency and accuracy direct driving device in plastic forming equipment.

  7. Vibration properties of a rotating piezoelectric energy harvesting device that experiences gyroscopic effects

    Science.gov (United States)

    Lu, Haohui; Chai, Tan; Cooley, Christopher G.

    2018-03-01

    This study investigates the vibration of a rotating piezoelectric device that consists of a proof mass that is supported by elastic structures with piezoelectric layers. Vibration of the proof mass causes deformation in the piezoelectric structures and voltages to power the electrical loads. The coupled electromechanical equations of motion are derived using Newtonian mechanics and Kirchhoff's circuit laws. The free vibration behavior is investigated for devices with identical (tuned) and nonidentical (mistuned) piezoelectric support structures and electrical loads. These devices have complex-valued, speed-dependent eigenvalues and eigenvectors as a result of gyroscopic effects caused by their constant rotation. The characteristics of the complex-valued eigensolutions are related to physical behavior of the device's vibration. The free vibration behaviors differ significantly for tuned and mistuned devices. Due to gyroscopic effects, the proof mass in the tuned device vibrates in either forward or backward decaying circular orbits in single-mode free response. This is proven analytically for all tuned devices, regardless of the device's specific parameters or operating speed. For mistuned devices, the proof mass has decaying elliptical forward and backward orbits. The eigenvalues are shown to be sensitive to changes in the electrical load resistances. Closed-form solutions for the eigenvalues are derived for open and close circuits. At high rotation speeds these devices experience critical speeds and instability.

  8. Is medical linac suitable for high-precision stereotactic irradiation? Investigations in geometrical accuracies of gantry and couch

    International Nuclear Information System (INIS)

    Kunieda, Etsuo; Kitamura, Masayuki; Kawaguchi, Osamu; Shigematsu, Naoyuki; Ando, Yutaka; Kubo, Atsushi; Ohira, Takayuki; Tonai, Takenori; Kawase, Takeshi.

    1998-01-01

    Linac-based radiosurgery has many advantages over the gamma knife, including low initial cost and no need of source replacement. On the other hand, most of the medical linacs currently in use were not originally designed to be applied for radiosurgery, and, therefore, careful quality assurance programs are required. In the gantry-head of a linac, a small CCD video camera is mounted in a position optically identical to that of the x-ray source. The video signal from the camera was digitalized to be evaluated for geometrical errors. A metal ball fixed to the stereotactic base frame via XYZ-sliding rods was used as a simulated target. Displacements of the target from the isocenter were measured during rotation of the gantry. Displacements in the gantry-rotation plane were satisfactorily small, while those perpendicular to it were maximal at gantry position angles of 0deg and 180deg. This error might be caused by gravitational vending of the heavy gantry head. Although other major errors of the linac were within one millimeter, the center of coach rotation around the isocenter did not coincide with the center of gantry rotation, probably owing to gravitational vending. Special care should be taken when very small collimators are employed. (author)

  9. Device for sealing a rotating plug in a nuclear reactor

    International Nuclear Information System (INIS)

    Brandstetter, R.

    1975-01-01

    The invention relates to the sealing of a rotating plug in a nuclear reactor. The sealing arrangement comprises a friction track which is formed along the periphery of the top of a ring mounted on a stationary element. An annular base coaxial with the plug is secured in sealing-tight manner to the stationary bearing around the ring and the track by means of a seal which rests on the annular base and also on the friction track of the ring and which comprises at least one friction ring and a clamping spring ring. The seal is clamped against the friction track to retractable clamping means when the plug is stationary, the retractable clamping means being carried by a ring secured to the first-mentioned ring. (U.S.)

  10. SU-G-TeP2-04: Comprehensive Machine Isocenter Evaluation with Separation of Gantry, Collimator, and Table Variables

    Energy Technology Data Exchange (ETDEWEB)

    Hancock, S [Southeast Missouri Hospital, Cape Girardeau, MO (United States); Clements, C [Radiological Imaging Technology, Colorado Springs, CO (United States); Hyer, D; Nixon, E [University Of Iowa, Iowa City, IA (United States); Martin, E; Wang, B [University of Louisville, Louisville, KY (United States); Jani, S [Sharp Memorial Hospital, San Diego, CA (United States); Gossman, M [Tri-State Regional Cancer Center, Ashland, KY (United States)

    2016-06-15

    Purpose: To develop and demonstrate application of a method that characterizes deviation of linac x-ray beams from the centroid of the volumetric radiation isocenter as a function of gantry, collimator, and table variables. Methods: A set of Winston-Lutz ball-bearing images was used to determine the gantry radiation isocenter as the midrange of deviation values resulting from gantry and collimator rotation. Also determined were displacement of table axis from gantry isocenter and recommended table axis adjustment. The method, previously reported, has been extended to include the effect of collimator walkout by obtaining measurements with 0 and 180 degree collimator rotation for each gantry angle. Twelve images were used to characterize the volumetric isocenter for the full range of available gantry, collimator, and table rotations. Results: Three Varian True Beam, two Elekta Infinity and four Versa HD linacs at five institutions were tested using identical methodology. Varian linacs exhibited substantially less deviation due to head sag than Elekta linacs (0.4 mm vs. 1.2 mm on average). One linac from each manufacturer had additional isocenter deviation of 0.3 to 0.4 mm due to jaw instability with gantry and collimator rotation. For all linacs, the achievable isocenter tolerance was dependent on adjustment of collimator position offset, transverse position steering, and alignment of the table axis with gantry isocenter, facilitated by these test results. The pattern and magnitude of table axis wobble vs. table angle was reproducible and unique to each machine. Conclusion: This new method provides a comprehensive set of isocenter deviation values including all variables. It effectively facilitates minimization of deviation between beam center and target (ball-bearing) position. This method was used to quantify the effect of jaw instability on isocenter deviation and to identify the offending jaw. The test is suitable for incorporation into a routine machine QA

  11. P-shaped Coiled Stator Ultrasound Motor for Rotating Intravascular Surgery Device

    Directory of Open Access Journals (Sweden)

    Toshinobu ABE

    2015-01-01

    Full Text Available The primary focus of this paper is the development of an ultra-miniature ultrasound motor for use in the human blood vessel. Since the size of the drive source for rotating the atherectomy device and intravascular ultrasonography system are large currently in practical use, it is installed outside the body, and the rotational power for the atherectomy device and intravascular ultrasonography system are transmitted through the long tortuous blood vessel. Such systems suffer from the problem that the rotation becomes non-uniform, and the problem that the available time is limited. We have therefore developed a P-shaped coiled stator ultrasound motor as a miniature ultrasound motor for rotating the ultrasound sensor for use in blood vessels in order to solve these problems. In this paper, we describe measurement of the torque, revolution speed, output power, efficiency, and particle motion on acoustic waveguide of the P-shaped coiled stator ultrasound motor.

  12. The effect of restraints type on the generated stresses in gantry crane beam

    Directory of Open Access Journals (Sweden)

    Sowa Leszek

    2018-01-01

    Full Text Available This paper includes an analysis of the mechanical phenomena in the gantry crane beam, because the cranes are currently one of the most common devices for the transporting loads. Designing modern mechanical structures is a complex task that requires the use of appropriate tools. Such a modern tool is the numerical simulation, which uses different numerical methods. One of the best known methods is the finite element method, also used here. Simulations are limited to analysis of the strength of the gantry crane beam that was the loaded of the force load movement along its length. The numerical analysis was made to the gantry crane beam which cross-section was an I-beam and ends were fixed in different ways. As the result of numerical calculations, the stresses and displacements of the structure of gantry were obtained. The influence of the restraints type and changing the loading force position on generate the Huber-Misses stress in the gantry crane beam was estimated. The aim was to ensure that the maximum equivalent stress generated in the gantry crane beam was less than the strength of material, because then the construction is safe.

  13. Noncontacting device to indicate deflection of turbopump internal rotating parts

    Science.gov (United States)

    Hamilton, D. B.; Grieser, D. R.; Plummer, A. M.; Ensminger, D.; Saccacio, E. J.

    1972-01-01

    Phase 2 (development) which was concluded for the ultrasonic Doppler device and the light-pipe-reflectance device is reported. An ultrasonic Doppler breadboard system was assembled which accurately measured runout in the J-2 LOX pump impeller during operation. The transducer was mounted on the outside of the pump volute using a C-clamp. Vibration was measured by conducting the ultrasonic wave through the volute housing and through the fluid in the volute to the impeller surface. The impeller vibration was also measured accurately using the light-pipe probe mounted in an elastomeric-gland fitting in the pump case. A special epoxy resin developed for cryogenic applications was forced into the end of the fiber-optic probe to retain the fibers. Subsequently, the probe suffered no damage after simultaneous exposure to 2150 psi and 77 F. Preliminary flash X-radiographs were taken of the turbine wheel and the shaft-bearing-seal assembly, using a 2-megavolt X-ray unit. Reasonable resolution and contrast was obtained. A fast-neutron detector was fabricated and sensitivity was measured. The results demonstrated that the technique is feasible for integrated-time measurements requiring, perhaps, 240 revolutions to obtain sufficient exposure at 35,000 rpm. The experimental verification plans are included.

  14. Vibrot, a simple device for the conversion of vibration into rotation mediated by friction: preliminary evaluation.

    Directory of Open Access Journals (Sweden)

    Ernesto Altshuler

    Full Text Available While "vibrational noise" induced by rotating components of machinery is a common problem constantly faced by engineers, the controlled conversion of translational into rotational motion or vice-versa is a desirable goal in many scenarios ranging from internal combustion engines to ultrasonic motors. In this work, we describe the underlying physics after isolating a single degree of freedom, focusing on devices that convert a vibration along the vertical axis into a rotation around this axis. A typical Vibrot (as we label these devices consists of a rigid body with three or more cantilevered elastic legs attached to its bottom at an angle. We show that these legs are capable of transforming vibration into rotation by a "ratchet effect", which is caused by the anisotropic stick-slip-flight motion of the leg tips against the ground. Drawing an analogy with the Froude number used to classify the locomotion dynamics of legged animals, we discuss the walking regime of these robots. We are able to control the rotation frequency of the Vibrot by manipulating the shaking amplitude, frequency or waveform. Furthermore, we have been able to excite Vibrots with acoustic waves, which allows speculating about the possibility of reducing the size of the devices so they can perform tasks into the human body, excited by ultrasound waves from the outside.

  15. Universal crystal cooling device for precession cameras, rotation cameras and diffractometers

    International Nuclear Information System (INIS)

    Hajdu, J.; McLaughlin, P.J.; Helliwell, J.R.; Sheldon, J.; Thompson, A.W.

    1985-01-01

    A versatile crystal cooling device is described for macromolecular crystallographic applications in the 290 to 80 K temperature range. It utilizes a fluctuation-free cold-nitrogen-gas supply, an insulated Mylar crystal cooling chamber and a universal ball joint, which connects the cooling chamber to the goniometer head and the crystal. The ball joint is a novel feature over all previous designs. As a result, the device can be used on various rotation cameras, precession cameras and diffractometers. The lubrication of the interconnecting parts with graphite allows the cooling chamber to remain stationary while the crystal and goniometer rotate. The construction allows for 360 0 rotation of the crystal around the goniometer axis and permits any settings on the arcs and slides of the goniometer head (even if working at 80 K). There are no blind regions associated with the frame holding the chamber. Alternatively, the interconnecting ball joint can be tightened and fixed. This results in a set up similar to the construction described by Bartunik and Schubert where the cooling chamber rotates with the crystal. The flexibility of the systems allows for the use of the device on most cameras or diffractometers. THis device has been installed at the protein crystallographic stations of the Synchrotron Radiation Source at Daresbury Laboratory and in the Laboratory of Molecular Biophysics, Oxford. Several data sets have been collected with processing statistics typical of data collected without a cooling chamber. Tests using the full white beam of the synchrotron also look promising. (orig./BHO)

  16. Rotation gate for a three-level superconducting quantum interference device qubit with resonant interaction

    International Nuclear Information System (INIS)

    Yang, C.-P.; Han Siyuan

    2006-01-01

    We show a way to realize an arbitrary rotation gate in a three-level superconducting quantum interference device (SQUID) qubit using resonant interaction. In this approach, the two logical states of the qubit are represented by the two lowest levels of the SQUID and a higher-energy intermediate level is utilized for the gate manipulation. By considering spontaneous decay from the intermediate level during the gate operation, we present a formula for calculating average fidelity over all possible initial states. Finally, based on realistic system parameters, we show that an arbitrary rotation gate can be achieved with a high fidelity in a SQUID

  17. Examination into the maximum rotational frequency for an in-plane switched active waveplate device

    International Nuclear Information System (INIS)

    Davidson, A J; Elston, S J; Raynes, E P

    2005-01-01

    An examination of an active waveplate device using a one-dimensional model, giving numerical and analytical results, is presented. The model calculates the director and twist configuration by minimizing the free energy of the system with simple homeotropic boundary conditions. The effect of varying the in-plane electric field in both magnitude and direction is examined, and it is shown that the twist through the cell is constant in time as the field is rotated. As the electric field is rotated, the director field lags behind by an angle which increases as the frequency of the electric field rotation increases. When this angle reaches approximately π/4 the director field no longer follows the electric field in a uniform way. Using mathematical analysis it is shown that the conditions on which the director profile will fail to follow the rotating electric field depend on the frequency of electric field rotation, the magnitude of the electric field, the dielectric anisotropy and the viscosity of the liquid crystal

  18. Design and Experimental Characterization of a Vibration Energy Harvesting Device for Rotational Systems

    Directory of Open Access Journals (Sweden)

    Lutao Yan

    2013-01-01

    Full Text Available This paper presents a new vibration based electromagnetic power generator to transfer energy from stationary to rotating equipment, which can be a new attempt to substitute slip ring in rotational systems. The natural frequencies and modes are simulated in order to have a maximum and steady power output from the device. Parameters such as piezoelectric disk location and relative motion direction of the magnet are theoretically and experimentally analyzed. The results show that the position that is close to the fixed end of the cantilever and the relative motion along the long side gives higher power output. Moreover, the capability of the energy harvester to extract power from lower energy environment is experimentally validated. The voltage and power output are measured at different excitation frequencies.

  19. SU-E-T-130: Are Proton Gantries Needed? An Analysis of 4332 Patient Proton Gantry Treatment Plans From the Past 10 Years

    International Nuclear Information System (INIS)

    Yan, S; Lu, H; Flanz, J; Depauw, N; Adams, J; Gorissen, BL; Wang, Y; Daartz, J; Bortfeld, T

    2015-01-01

    Purpose: To ascertain the necessity of a proton gantry, as compared to the feasibility of using a horizontal fixed proton beam-line for treatment with advanced technology. Methods: To calculate the percentage of patients that can be treated with a horizontal fixed beam-line instead of a gantry, we analyze the distributions of beam orientations of our proton gantry patients treated over the past 10 years. We identify three horizontal fixed beam geometries (FIXED, BEND and MOVE) with the patient in lying and/or sitting positions. The FIXED geometry includes only table/chair rotations and translations. In BEND, the beam can be bent up/down for up to 20 degrees. MOVE allows for patient head/body angle adjustment. Based on the analysis, we select eight patients whose plan involves beams which are still challenging to achieve with a horizontal fixed beam. These beams are removed in the pencil beam scanning (PBS) plan optimized for the fixed beam-line (PBS-fix). We generate non-coplanar PBS-gantry plans for comparison, and perform a robustness analysis. Results: The percentage of patients with head-and-neck/brain tumors that can be treated with horizontal fixed beam is 44% in FIXED, 70% in 20-degrees BEND, and 100% in 90-degrees MOVE. For torso regions, 99% of the patients can be treated in 20-degree BEND. The target coverage is more homogeneous with PBS-fix plans compared to the clinical scattering treatment plans. The PBS-fix plans reduce the mean dose to organs-at-risk by a factor of 1.1–28.5. PBS-gantry plans are as good as PBS-fix plans, sometimes marginally better. Conclusion: The majority of the beam orientations can be realized with a horizontal fixed beam-line. Challenging non-coplanar beams can be eliminated with PBS delivery. Clinical implementation of the proposed fixed beam-line requires use of robotic patient positioning, further developments in immobilization, and image guidance. However, our results suggest that fixed beam-lines can be as effective as

  20. SU-E-T-130: Are Proton Gantries Needed? An Analysis of 4332 Patient Proton Gantry Treatment Plans From the Past 10 Years

    Energy Technology Data Exchange (ETDEWEB)

    Yan, S; Lu, H; Flanz, J; Depauw, N; Adams, J; Gorissen, BL; Wang, Y; Daartz, J; Bortfeld, T [Massachusetts General Hospital, Boston, MA (United States)

    2015-06-15

    Purpose: To ascertain the necessity of a proton gantry, as compared to the feasibility of using a horizontal fixed proton beam-line for treatment with advanced technology. Methods: To calculate the percentage of patients that can be treated with a horizontal fixed beam-line instead of a gantry, we analyze the distributions of beam orientations of our proton gantry patients treated over the past 10 years. We identify three horizontal fixed beam geometries (FIXED, BEND and MOVE) with the patient in lying and/or sitting positions. The FIXED geometry includes only table/chair rotations and translations. In BEND, the beam can be bent up/down for up to 20 degrees. MOVE allows for patient head/body angle adjustment. Based on the analysis, we select eight patients whose plan involves beams which are still challenging to achieve with a horizontal fixed beam. These beams are removed in the pencil beam scanning (PBS) plan optimized for the fixed beam-line (PBS-fix). We generate non-coplanar PBS-gantry plans for comparison, and perform a robustness analysis. Results: The percentage of patients with head-and-neck/brain tumors that can be treated with horizontal fixed beam is 44% in FIXED, 70% in 20-degrees BEND, and 100% in 90-degrees MOVE. For torso regions, 99% of the patients can be treated in 20-degree BEND. The target coverage is more homogeneous with PBS-fix plans compared to the clinical scattering treatment plans. The PBS-fix plans reduce the mean dose to organs-at-risk by a factor of 1.1–28.5. PBS-gantry plans are as good as PBS-fix plans, sometimes marginally better. Conclusion: The majority of the beam orientations can be realized with a horizontal fixed beam-line. Challenging non-coplanar beams can be eliminated with PBS delivery. Clinical implementation of the proposed fixed beam-line requires use of robotic patient positioning, further developments in immobilization, and image guidance. However, our results suggest that fixed beam-lines can be as effective as

  1. Commissioning of a proton gantry equipped with dual x-ray imagers and a robotic patient positioner, and evaluation of the accuracy of single-beam image registration for this system

    International Nuclear Information System (INIS)

    Wang, Ning; Ghebremedhin, Abiel; Patyal, Baldev

    2015-01-01

    Purpose: To check the accuracy of a gantry equipped with dual x-ray imagers and a robotic patient positioner for proton radiotherapy, and to evaluate the accuracy and feasibility of single-beam registration using the robotic positioner. Methods: One of the proton treatment rooms at their institution was upgraded to include a robotic patient positioner (couch) with 6 degrees of freedom and dual orthogonal kilovoltage x-ray imaging panels. The wander of the proton beam central axis, the wander of the beamline, and the orthogonal image panel crosswires from the gantry isocenter were measured for different gantry angles. The couch movement accuracy and couch wander from the gantry isocenter were measured for couch loadings of 50–300 lb with couch rotations from 0° to ±90°. The combined accuracy of the gantry, couch, and imagers was checked using a custom-made 30 × 30 × 30 cm 3 Styrofoam phantom with beekleys embedded in it. A treatment in this room can be set up and registered at a setup field location, then moved precisely to any other treatment location without requiring additional image registration. The accuracy of the single-beam registration strategy was checked for treatments containing multiple beams with different combinations of gantry angles, couch yaws, and beam locations. Results: The proton beam central axis wander from the gantry isocenter was within 0.5 mm with gantry rotations in both clockwise (CW) and counterclockwise (CCW) directions. The maximum wander of the beamline and orthogonal imager crosswire centers from the gantry isocenter were within 0.5 and 0.8 mm, respectively, with the gantry rotations in CW and CCW directions. Vertical and horizontal couch wanders from the gantry isocenter were within 0.4 and 1.3 mm, respectively, for couch yaw from 0° to ±90°. For a treatment with multiple beams with different gantry angles, couch yaws, and beam locations, the measured displacements of treatment beam locations from the one based on the

  2. Magnetic field devices for neutron spin transport and manipulation in precise neutron spin rotation measurements

    Energy Technology Data Exchange (ETDEWEB)

    Maldonado-Velázquez, M. [Posgrado en Ciencias Físicas, Universidad Nacional Autónoma de México, 04510 (Mexico); Barrón-Palos, L., E-mail: libertad@fisica.unam.mx [Instituto de Física, Universidad Nacional Autónoma de México, Apartado Postal 20-364, 01000 (Mexico); Crawford, C. [University of Kentucky, Lexington, KY 40506 (United States); Snow, W.M. [Indiana University, Bloomington, IN 47405 (United States)

    2017-05-11

    The neutron spin is a critical degree of freedom for many precision measurements using low-energy neutrons. Fundamental symmetries and interactions can be studied using polarized neutrons. Parity-violation (PV) in the hadronic weak interaction and the search for exotic forces that depend on the relative spin and velocity, are two questions of fundamental physics that can be studied via the neutron spin rotations that arise from the interaction of polarized cold neutrons and unpolarized matter. The Neutron Spin Rotation (NSR) collaboration developed a neutron polarimeter, capable of determining neutron spin rotations of the order of 10{sup −7} rad per meter of traversed material. This paper describes two key components of the NSR apparatus, responsible for the transport and manipulation of the spin of the neutrons before and after the target region, which is surrounded by magnetic shielding and where residual magnetic fields need to be below 100 μG. These magnetic field devices, called input and output coils, provide the magnetic field for adiabatic transport of the neutron spin in the regions outside the magnetic shielding while producing a sharp nonadiabatic transition of the neutron spin when entering/exiting the low-magnetic-field region. In addition, the coils are self contained, forcing the return magnetic flux into a compact region of space to minimize fringe fields outside. The design of the input and output coils is based on the magnetic scalar potential method.

  3. Variable speed control in wells turbine-based oscillating water column devices: optimum rotational speed

    Science.gov (United States)

    Lekube, J.; Garrido, A. J.; Garrido, I.

    2018-03-01

    The effects of climate change and global warming reveal the need to find alternative sources of clean energy. In this sense, wave energy power plants, and in particular Oscillating Water Column (OWC) devices, offer a huge potential of energy harnessing. Nevertheless, the conversion systems have not reached a commercially mature stage yet so as to compete with conventional power plants. At this point, the use of new control methods over the existing technology arises as a doable way to improve the efficiency of the system. Due to the non-uniform response that the turbine shows to the rotational speed variation, the speed control of the turbo-generator may offer a feasible solution for efficiency improvement during the energy conversion. In this context, a novel speed control approach for OWC systems is presented in this paper, demonstrating its goodness and affording promising results when particularized to the Mutriku’s wave power plant.

  4. Rotational population patterns and searches for the nuclear SQUID (Superconducting Quantum Interference Device)

    International Nuclear Information System (INIS)

    Canto, L.F.; Donangelo, R.J.; Farhan, A.R.; Guidry, M.W.; Rasmussen, J.O.; Ring, P.; Stoyer, M.A.

    1989-11-01

    This paper presents new theoretical results for rotational population patterns in the nuclear SQUID effect. (The term nuclear SQUID is in analogy to the solid-state Superconducting Quantum Interference Devices.) The SQUID effect is an interesting new twist to an old quest to understand Coriolis anti-pairing (CAP) effects in nuclear rotational bands. Two-neutron transfer reaction cross sections among high-spin states have long been touted as more specific CAP probes than other nuclear properties. Heavy projectiles like Sn or Pb generally are recommended to pump the deformed nucleus to as high spin as possible for transfer. The interference and sign reversal of 2n transfer amplitudes at high spin, as predicted in the early SQUID work imposes the difficult requirement of Coulomb pumping to near back-bending spins at closest approach. For Pb on rare earths we find a dramatic departure from sudden-approximation, so that the population depression occurs as low as final spin 10h. 14 refs., 8 figs

  5. Moving gantry method for electron beam dose profile measurement at extended source-to-surface distances.

    Science.gov (United States)

    Fekete, Gábor; Fodor, Emese; Pesznyák, Csilla

    2015-03-08

    A novel method has been put forward for very large electron beam profile measurement. With this method, absorbed dose profiles can be measured at any depth in a solid phantom for total skin electron therapy. Electron beam dose profiles were collected with two different methods. Profile measurements were performed at 0.2 and 1.2 cm depths with a parallel plate and a thimble chamber, respectively. 108cm × 108 cm and 45 cm × 45 cm projected size electron beams were scanned by vertically moving phantom and detector at 300 cm source-to-surface distance with 90° and 270° gantry angles. The profiles collected this way were used as reference. Afterwards, the phantom was fixed on the central axis and the gantry was rotated with certain angular steps. After applying correction for the different source-to-detector distances and incidence of angle, the profiles measured in the two different setups were compared. Correction formalism has been developed. The agreement between the cross profiles taken at the depth of maximum dose with the 'classical' scanning and with the new moving gantry method was better than 0.5 % in the measuring range from zero to 71.9 cm. Inverse square and attenuation corrections had to be applied. The profiles measured with the parallel plate chamber agree better than 1%, except for the penumbra region, where the maximum difference is 1.5%. With the moving gantry method, very large electron field profiles can be measured at any depth in a solid phantom with high accuracy and reproducibility and with much less time per step. No special instrumentation is needed. The method can be used for commissioning of very large electron beams for computer-assisted treatment planning, for designing beam modifiers to improve dose uniformity, and for verification of computed dose profiles.

  6. Can Rotational Atherectomy Cause Thermal Tissue Damage? A Study of the Potential Heating and Thermal Tissue Effects of a Rotational Atherectomy Device

    International Nuclear Information System (INIS)

    Gehani, Abdurrazzak A.; Rees, Michael R.

    1998-01-01

    Purpose: Thermal tissue damage (TTD) is customarily associated with some lasers. The thermal potential of rotational atherectomy (RA) devices is unknown. We investigated the temperature profile and potential TTD as well as the value of fluid flushing of an RA device. Methods: We used a high-resolution infrared imaging system that can detect changes as small as 0.1 deg. C to measure the temperature changes at the tip of a fast RA device with and without fluid flushing. To assess TTD, segments of porcine aorta were subjected to the rotating tip under controlled conditions, stained by a special histochemical stain (picrisirius red) and examined under normal and polarized light microscopy. Results: There was significant heating of the rotating cam. The mean 'peak' temperature rise was 52.8 ± 16.9 deg. C. This was related to rotational speed; thus the 'peak' temperature rise was 88.3 ± 12.6 deg. C at 80,000 rpm and 17.3 ± 3.8 deg. C at 20,000 rpm (p < 0.001, t-test). Fluid flushing at 18 ml/min reduced, but did not abolish, heating of the device (11.8 ± 2.9 deg. C). A crater was observed in all segments exposed to the rotating tip. The following features were most notable: (i) A zone of 'thermal' tissue damage extended radially from the crater reaching adventitia in some sections, especially at high speeds. This zone showed markedly reduced or absent birefringence. (ii) Fluid flushing of the catheter reduced the above changes but increased the incidence and extent of dissections in the media, especially when combined with high atherectomy speeds. (iii) These changes were observed in five of six specimens exposed to RA without flushing, but in only one of six with flushing (p < 0.05). (iv) None of the above changes was seen in control segments. Conclusion: RA is capable of generating significant heat and potential TTD. Fluid flushing reduced heating and TTD. These findings warrant further studies in vivo, and may influence the design of atherectomy devices

  7. VMATc: VMAT with constant gantry speed and dose rate

    International Nuclear Information System (INIS)

    Peng, Fei; Romeijn, H Edwin; Epelman, Marina A; Jiang, Steve B

    2015-01-01

    This article considers the treatment plan optimization problem for Volumetric Modulated Arc Therapy (VMAT) with constant gantry speed and dose rate (VMATc). In particular, we consider the simultaneous optimization of multi-leaf collimator leaf positions and a constant gantry speed and dose rate. We propose a heuristic framework for (approximately) solving this optimization problem that is based on hierarchical decomposition. Specifically, an iterative algorithm is used to heuristically optimize dose rate and gantry speed selection, where at every iteration a leaf position optimization subproblem is solved, also heuristically, to find a high-quality plan corresponding to a given dose rate and gantry speed. We apply our framework to clinical patient cases, and compare the resulting VMATc plans to idealized IMRT, as well as full VMAT plans. Our results suggest that VMATc is capable of producing treatment plans of comparable quality to VMAT, albeit at the expense of long computation time and generally higher total monitor units. (paper)

  8. Force Outputs during Squats Performed Using a Rotational Inertia Device under Stable versus Unstable Conditions with Different Loads.

    Science.gov (United States)

    Vázquez-Guerrero, Jairo; Moras, Gerard; Baeza, Jennifer; Rodríguez-Jiménez, Sergio

    2016-01-01

    The purpose of the study was to compare the force outputs achieved during a squat exercise using a rotational inertia device in stable versus unstable conditions with different loads and in concentric and eccentric phases. Thirteen male athletes (mean ± SD: age 23.7 ± 3.0 years, height 1.80 ± 0.08 m, body mass 77.4 ± 7.9 kg) were assessed while squatting, performing one set of three repetitions with four different loads under stable and unstable conditions at maximum concentric effort. Overall, there were no significant differences between the stable and unstable conditions at each of the loads for any of the dependent variables. Mean force showed significant differences between some of the loads in stable and unstable conditions (P inertia device allowed the generation of similar force outputs under stable and unstable conditions at each of the four loads. The study also provides empirical evidence of the different force outputs achieved by adjusting load conditions on the rotational inertia device when performing squats, especially in the case of peak force. Concentric force outputs were significantly higher than eccentric outputs, except for peak force under both conditions. These findings support the use of the rotational inertia device to train the squatting exercise under unstable conditions for strength and conditioning trainers. The device could also be included in injury prevention programs for muscle lesions and ankle and knee joint injuries.

  9. Force production during squats performed with a rotational resistance device under stable versus unstable conditions.

    Science.gov (United States)

    Moras, Gerard; Vázquez-Guerrero, Jairo

    2015-11-01

    [Purpose] Force production during a squat action on a rotational resistance device (RRD) under stable and unstable conditions. [Subjects and Methods] Twenty-one healthy males were asked to perform six sets of six repetitions of squats on an RRD on either stable or unstable surfaces. The stable and unstable sets were performed on different days. Muscular outputs were obtained from a linear encoder and a strain gauge fixed to a vest. [Results] Overall, the results showed no significant differences for any of the dependent variables across exercise modes. Forcemean outputs were higher in the concentric phase than in the eccentric phase for each condition, but there were no differences in velocity, time or displacement. The forcepeak was similar in the eccentric and concentric phases of movement under both stable and unstable conditions. There were no significant differences in forcemean between sets per condition or between conditions. [Conclusion] These results suggest that performing squats with a RRD achieves similar forcemean and forcepeak under stable and unstable conditions. The forcepeak produced is also similar in concentric and eccentric phases.

  10. Wireless Synchronization of a Multi-Pinhole Small Animal SPECT Collimation Device With a Clinical Scanner

    Science.gov (United States)

    DiFilippo, Frank P.; Patel, Sagar

    2009-06-01

    A multi-pinhole collimation device for small animal single photon emission computed tomography (SPECT) uses the gamma camera detectors of a standard clinical SPECT scanner. The collimator and animal bed move independently of the detectors, and therefore their motions must be synchronized. One approach is manual triggering of the SPECT acquisition simultaneously with a programmed motion sequence for the device. However, some data blurring and loss of image quality result, and true electronic synchronization is preferred. An off-the-shelf digital gyroscope with integrated Bluetooth interface provides a wireless solution to device synchronization. The sensor attaches to the SPECT gantry and reports its rotational speed to a notebook computer controlling the device. Software processes the rotation data in real-time, averaging the signal and issuing triggers while compensating for baseline drift. Motion commands are sent to the collimation device with minimal delay, within approximately 0.5 second of the start of SPECT gantry rotation. Test scans of a point source demonstrate an increase in true counts and a reduction in background counts compared to manual synchronization. The wireless rotation sensor provides robust synchronization of the collimation device with the clinical SPECT scanner and enhances image quality.

  11. EMPLACEMENT GANTRY ITS STANDARDS IDENTIFICATION STUDY

    International Nuclear Information System (INIS)

    Voegele, M.

    2005-01-01

    To date, the project has established ITS performance requirements for SSCs based on identification and categorization of event sequences that may result in a radiological release. These performance requirements are defined within the NSDB. Further, SSCs credited with performing safe functions are classified as ITS. In turn, perform confirmation for these SSCs is sought through the use of consensus code and standards. The purpose of this study is to identify applicable codes and standards for the WP Emplacement Gantry ITS SSCs. Further, this study will form the basis for selection and the extent of applicability of each code and standard. This study is based on the design development completed for LA only. Accordingly, identification of ITS SSCs beyond those defined within the NSDB are based on designs that may be subject to further development during detail design. Furthermore, several design alternatives may still be under consideration to satisfy certain safety functions, and that final selection will not be determined until further design development has occurred. Therefore, for completeness, throughout this study alternative designs currently under considered will be discussed. Further, the results of this study will be subject to evaluation as part of a follow-on GAP analysis study. Based on the results of this study the GAP analysis will evaluate each code and standard to ensure each ITS performance requirement is fully satisfied. When a performance requirement is not fully satisfied a ''gap'' is highlighted. Thereafter, the study will identify supplemental requirements to augment the code or standard to meet performance requirements. Further, the GAP analysis will identify non-standard areas of the design that will be subject to a Development Plan. Non-standard components and non-standard design configurations are defined as areas of the design that do not follow standard industry practices or codes and standards. Whereby, performance confirmation cannot be

  12. Force Outputs during Squats Performed Using a Rotational Inertia Device under Stable versus Unstable Conditions with Different Loads.

    Directory of Open Access Journals (Sweden)

    Jairo Vázquez-Guerrero

    Full Text Available The purpose of the study was to compare the force outputs achieved during a squat exercise using a rotational inertia device in stable versus unstable conditions with different loads and in concentric and eccentric phases. Thirteen male athletes (mean ± SD: age 23.7 ± 3.0 years, height 1.80 ± 0.08 m, body mass 77.4 ± 7.9 kg were assessed while squatting, performing one set of three repetitions with four different loads under stable and unstable conditions at maximum concentric effort. Overall, there were no significant differences between the stable and unstable conditions at each of the loads for any of the dependent variables. Mean force showed significant differences between some of the loads in stable and unstable conditions (P < 0.010 and peak force output differed between all loads for each condition (P < 0.045. Mean force outputs were greater in the concentric than in the eccentric phase under both conditions and with all loads (P < 0.001. There were no significant differences in peak force between concentric and eccentric phases at any load in either stable or unstable conditions. In conclusion, squatting with a rotational inertia device allowed the generation of similar force outputs under stable and unstable conditions at each of the four loads. The study also provides empirical evidence of the different force outputs achieved by adjusting load conditions on the rotational inertia device when performing squats, especially in the case of peak force. Concentric force outputs were significantly higher than eccentric outputs, except for peak force under both conditions. These findings support the use of the rotational inertia device to train the squatting exercise under unstable conditions for strength and conditioning trainers. The device could also be included in injury prevention programs for muscle lesions and ankle and knee joint injuries.

  13. Evaluation of a dual-room sliding gantry CT concept for workflow optimisation in polytrauma and regular in- and outpatient management

    Energy Technology Data Exchange (ETDEWEB)

    Frellesen, Claudia; Boettcher, Marie; Wichmann, Julian L.; Drieske, Martina; Kerl, J. Matthias; Lehnert, Thomas [Department of Diagnostic and Interventional Radiology, Clinic of the Goethe University, Frankfurt (Germany); Nau, Christoph; Geiger, Emmanuel; Wutzler, Sebastian [Department of Trauma, Reconstructive and Hand Surgery, Clinic of the Goethe University, Frankfurt (Germany); Ackermann, Hanns [Department of Biostatistics and Mathematical Modelling, Clinic of the Goethe University, Theodor-Stern-Kai 7, 60590 Frankfurt (Germany); Vogl, Thomas J. [Department of Diagnostic and Interventional Radiology, Clinic of the Goethe University, Frankfurt (Germany); Bauer, Ralf W., E-mail: ralfwbauer@aol.com [Department of Diagnostic and Interventional Radiology, Clinic of the Goethe University, Frankfurt (Germany)

    2015-01-15

    Highlights: • A sliding gantry trauma room CT solution facilitates significantly faster polytrauma management. • Faster and more efficient resumption of regularly scheduled patients due to a two room solution is supported. • Sliding gantry CT achieves the same patient throughput as two separate conventional CT devices. - Abstract: Objectives: To reveal the impact on workflow from introducing a dual-room sliding gantry CT to the trauma room for polytrauma and regularly scheduled in- outpatients with regard to efficiency and degree of capacity utilisation. Materials and methods: Time analysis was performed for 30 polytrauma patients each in 2 different trauma room settings, the new trauma room comprising a sliding gantry CT, the old one a stationary single-room CT. Complete trauma room and diagnostic workup times were manually measured and compared for both groups. In a third scenario, the number of CT scans performed with one single sliding gantry CT and the two-room concept was compared to the number of CT scans performed on two separate regular CT units in a 5 days clinical routine sample. Results: Patients demographics and type of CT examinations were comparable for all patient groups. The median time from patient arrival in the trauma room until beginning of CT scanning was 6 min shorter for the sliding gantry CT group (21 vs.15 min). Sliding gantry CT embedded in a two-room solution achieved 252 CT scans in 5 working days, compared to 250 CT scans on two separate regular CT units with the same man power. Conclusions: Sliding gantry CT in the trauma room allows for significant time saving in the diagnostic workup of polytrauma patients and faster resumption of the regular in- outpatient's CT schedule is possible. With the same man power, the dual-room solution is able to generate the same throughput as two separate CT units.

  14. Evaluation of a dual-room sliding gantry CT concept for workflow optimisation in polytrauma and regular in- and outpatient management

    International Nuclear Information System (INIS)

    Frellesen, Claudia; Boettcher, Marie; Wichmann, Julian L.; Drieske, Martina; Kerl, J. Matthias; Lehnert, Thomas; Nau, Christoph; Geiger, Emmanuel; Wutzler, Sebastian; Ackermann, Hanns; Vogl, Thomas J.; Bauer, Ralf W.

    2015-01-01

    Highlights: • A sliding gantry trauma room CT solution facilitates significantly faster polytrauma management. • Faster and more efficient resumption of regularly scheduled patients due to a two room solution is supported. • Sliding gantry CT achieves the same patient throughput as two separate conventional CT devices. - Abstract: Objectives: To reveal the impact on workflow from introducing a dual-room sliding gantry CT to the trauma room for polytrauma and regularly scheduled in- outpatients with regard to efficiency and degree of capacity utilisation. Materials and methods: Time analysis was performed for 30 polytrauma patients each in 2 different trauma room settings, the new trauma room comprising a sliding gantry CT, the old one a stationary single-room CT. Complete trauma room and diagnostic workup times were manually measured and compared for both groups. In a third scenario, the number of CT scans performed with one single sliding gantry CT and the two-room concept was compared to the number of CT scans performed on two separate regular CT units in a 5 days clinical routine sample. Results: Patients demographics and type of CT examinations were comparable for all patient groups. The median time from patient arrival in the trauma room until beginning of CT scanning was 6 min shorter for the sliding gantry CT group (21 vs.15 min). Sliding gantry CT embedded in a two-room solution achieved 252 CT scans in 5 working days, compared to 250 CT scans on two separate regular CT units with the same man power. Conclusions: Sliding gantry CT in the trauma room allows for significant time saving in the diagnostic workup of polytrauma patients and faster resumption of the regular in- outpatient's CT schedule is possible. With the same man power, the dual-room solution is able to generate the same throughput as two separate CT units

  15. Long-term successful arthroscopic repair of large and massive rotator cuff tears with a functional and degradable reinforcement device.

    Science.gov (United States)

    Proctor, Christopher S

    2014-10-01

    Rotator cuff repair is a procedure with varying outcomes, and there has been subsequent interest in devices that reinforce the repair and enhance structural and functional outcomes. The objective of this study was to determine these outcomes for arthroscopic repair of large and massive rotator cuff tears augmented with a synthetic absorbable mesh designed specifically for reinforcement of tendon repair by imaging and clinical assessments. Consecutive arthroscopic repairs were performed on 18 patients with large to massive rotator cuff tears by use of a poly-l-lactic acid synthetic patch as a reinforcement device and fixation with 4 sutures. Patients were assessed preoperatively and at 6 months, 12 months, and a mean of 42 months after surgery by the American Shoulder and Elbow Surgeons (ASES) shoulder score to evaluate clinical performance and at 12 months by ultrasound to assess structural repair. Ultrasound showed that 15 of 18 patients had intact rotator cuff repair at 12 months; at 42 months, an additional patient had a failed repair. Patients showed improvement in the ASES shoulder score from 25 preoperatively to 71 at 12 months and 70 at 42 months after surgery. Patients with intact rotator cuff (n = 14) at 42 months had an ASES shoulder score of 82. The poly-l-lactic acid bioabsorbable patch designed specifically to reinforce the surgical repair of tendons supported successful repair of large to massive rotator cuff tears in 83% of patients at 12 months after surgery and 78% of patients at 42 months after surgery, with substantial functional improvement. Copyright © 2014 Journal of Shoulder and Elbow Surgery Board of Trustees. Published by Elsevier Inc. All rights reserved.

  16. Design of a rotational three-dimensional nonimaging device by a compensated two-dimensional design process.

    Science.gov (United States)

    Yang, Yi; Qian, Ke-Yuan; Luo, Yi

    2006-07-20

    A compensation process has been developed to design rotational three-dimensional (3D) nonimaging devices. By compensating the desired light distribution during a two-dimensional (2D) design process for an extended Lambertian source using a compensation coefficient, the meridian plane of a 3D device with good performance can be obtained. This method is suitable in many cases with fast calculation speed. Solutions to two kinds of optical design problems have been proposed, and the limitation of this compensated 2D design method is discussed.

  17. Carbon/proton therapy: A novel gantry design

    Directory of Open Access Journals (Sweden)

    D. Trbojevic

    2007-05-01

    Full Text Available A major expense and design challenge in carbon/proton cancer therapy machines are the isocentric gantries. The transport elements of the carbon/proton gantry are presently made of standard conducting dipoles. Because of their large weight, of the order of ∼100   tons, the total weight of the gantry with support structure is ∼600   tons. The novel gantry design that is described here is made of fixed field superconducting magnets, thus dramatically reducing magnet size and weight compared to conventional magnets. In addition, the magnetic field is constant throughout the whole energy region required for tumor treatment. Particles make very small orbit offsets, passing through the beam line. The beam line is built of combined-function dipoles such as a nonscaling fixed field alternating gradient (NS-FFAG structure. The very large momentum acceptance NS-FFAG comes from very strong focusing and very small dispersion. The NS-FFAG small magnets almost completely filled the beam line. They first make a quarter (or close to a quarter of an arc bending upward and an additional half of a circle beam line finishing so that the beam is pointed towards the patient. At the end of the gantry, additional magnets with a fast response are required to allow radial scanning and to provide the required position and spot size. The fixed field combined-function magnets for the carbon gantry could be made of superconducting magnets by using low temperature superconducting cable or by using high temperature superconductors.

  18. Dosimetry with the scanned proton beam on the PSI gantry

    International Nuclear Information System (INIS)

    Coray, A.; Pedroni, E.; Boehringer, T.; Lin, S.; Lomax, T.; Goitein, G.

    2002-01-01

    the precision and reproducibility of the dose application, independent of the gantry angle and the beam energy, will be presented. Before the first delivery of a new field to the patient, the dose distribution is checked by using a water phantom with an array of ionization chambers. A qualitative 3d check is done in addition through irradiation of a stack of films. The routine dosimetry with ionization chambers agree well with the expected dose from the therapy plan. The overall dose error for one year is ±0.03 Gy (S.D.). Some systematic effects, due to reactions in the range-shifter plates, have been found for small fields. Data to illustrate the dose verification will be shown. For our quality control and for the further refinement of the application technique we use a CCD dosimetry system. A scintillating screen is viewed by a CCD camera and a mirror. This 2d dosimetry device has an excellent position resolution and shows a very good reproducibility of the beam delivery. Further developments with the aim of realizing a 3D dosimetry system for the measurement of intensity modulated fields, which have a high spatial resolution, are under way. (author)

  19. Mechanical analysis and optimisation of large and highly-loaded bearing rollers For the "Riesenrad" Ion Gantry

    CERN Document Server

    Reimoser, S A

    2000-01-01

    A carbon ion gantry would allow the irradiation of cancer patients with carbon ions from any direction in space best suited for therapy. Till today, such a machine has not been built due to the expected size, mass and cost. A novel design, called "Riesenrad" ion gantry, promises to provide a competitive solution. The central part of the Riesenrad, which can rotate ± 90°, is supported (statically determinate) on pendular bearing units with two rollers each. High precision requirements for the structure rule out any plastic deformations in the area of contact. The present report describes the design of the highly-loaded rollers. In order to achieve a large contact area and a uniform distribution of contact stresses, a "barrel shape" for the rollers is proposed. An analysis using the finite element method (FEM) was performed to optimise the roller design, namely to establish the required crown roll (camber radius).

  20. The modification and the installation of WILLI detector as a rotatable device

    International Nuclear Information System (INIS)

    Brancus, I. M.; Vulpescu, B.; Badea, A. F.; Bozdog, H.; Duma, M.; Petcu, M.; Aiftimiei, C.; Mitrica, B.; Wentz, J.; Rebel, H.; Bercuci, A.; Mathes, H. J.

    2001-01-01

    The compact WILLI device, built in IFIN-HH Bucharest (44026'N latitude and 26004'E longitude) at a vertical cut-off rigidity of 5.6 GV, has been used for measurements of the charge ratio in the vertical atmospheric muon flux at momenta bellow 1GeV/c. In this low energy range the studies of muon charge ratio provide information useful for the discussion of the so-called atmospheric neutrino anomaly and for studies of atmospheric neutrino and antineutrino fluxes. The experimental method is based on the observation of the reduced effective lifetime of the negative muons, stopped in matter, as compared to the lifetime of positive muons. Avoiding the difficulties and the systematic errors of magnetic spectrometers, results are obtained with high accuracy, indicating a decrease of the muon charge ratio from 1.30 (at 0.87 GeV/c) to 1.15 (at 0.24 GeV/c). Super-Kamiokande and other experiments find that the ratio of muonic to electronic neutrinos is much smaller than the theoretical production, the effect depending on the incidence angle of the neutrinos. The interpretation of such so-called 'neutrino anomaly' in terms of neutrino oscillations is based on theoretical predictions of the fluxes of the neutrinos of different flavour, based on different hadronic interaction models. A possibility to verify such models is to compare the theoretical predictions for muon flux, which plays a key role in neutrino chain, with the experimental data. Thus, the ratio of the flux of positive to negative muons could provide a sensitive test for models regarding muon and neutrino production. The WILLI detector is a suitable instrument for further investigation of the modulation of muon charge ratio and the neutrino fluxes by the geomagnetic cut-off. The details about the geomagnetic influence have not been explored using the observation of muons with different angles of incidence in zenithal and azimuthal plane. The WILLI detector is now modified in a rotatable set-up, which will allow

  1. Roll and pitch set-up errors during volumetric modulated arc delivery: can adapting gantry and collimator angles compensate?

    Science.gov (United States)

    Hoffmans-Holtzer, Nienke A; Hoffmans, Daan; Dahele, Max; Slotman, Ben J; Verbakel, Wilko F A R

    2015-03-01

    The purpose of this work was to investigate whether adapting gantry and collimator angles can compensate for roll and pitch setup errors during volumetric modulated arc therapy (VMAT) delivery. Previously delivered clinical plans for locally advanced head-and-neck (H&N) cancer (n = 5), localized prostate cancer (n = 2), and whole brain with simultaneous integrated boost to 5 metastases (WB + 5M, n = 1) were used for this study. Known rigid rotations were introduced in the planning CT scans. To compensate for these, in-house software was used to adapt gantry and collimator angles in the plan. Doses to planning target volumes (PTV) and critical organs at risk (OAR) were calculated with and without compensation and compared with the original clinical plan. Measurements in the sagittal plane in a polystyrene phantom using radiochromic film were compared by gamma (γ) evaluation for 2 H&N cancer patients. For H&N plans, the introduction of 2°-roll and 3°-pitch rotations reduced mean PTV coverage from 98.7 to 96.3%. This improved to 98.1% with gantry and collimator compensation. For prostate plans respective figures were 98.4, 97.5, and 98.4%. For WB + 5M, compensation worked less well, especially for smaller volumes and volumes farther from the isocenter. Mean comparative γ evaluation (3%, 1 mm) between original and pitched plans resulted in 86% γ plan restored the mean comparison to 96% γ < 1. Preliminary data suggest that adapting gantry and collimator angles is a promising way to correct roll and pitch set-up errors of < 3° during VMAT for H&N and prostate cancer.

  2. New system for linear accelerator radiosurgery with a gantry-mounted video camera

    International Nuclear Information System (INIS)

    Kunieda, Etsuo; Kitamura, Masayuki; Kawaguchi, Osamu; Ohira, Takayuki; Ogawa, Kouichi; Ando, Yutaka; Nakamura, Kayoko; Kubo, Atsushi

    1998-01-01

    Purpose: We developed a positioning method that does not depend on the positioning mechanism originally annexed to the linac and investigated the positioning errors of the system. Methods and Materials: A small video camera was placed at a location optically identical to the linac x-ray source. A target pointer comprising a convex lens and bull's eye was attached to the arc of the Leksell stereotactic system so that the lens would form a virtual image of the bull's eye (virtual target) at the position of the center of the arc. The linac gantry and target pointer were placed at the side and top to adjust the arc center to the isocenter by referring the virtual target. Coincidence of the target and the isocenter could be confirmed in any combination of the couch and gantry rotation. In order to evaluate the accuracy of the positioning, a tungsten ball was attached to the stereotactic frame as a simulated target, which was repeatedly localized and repositioned to estimate the magnitude of the error. The center of the circular field defined by the collimator was marked on the film. Results: The differences between the marked centers of the circular field and the centers of the shadow of the simulated target were less than 0.3 mm

  3. A novel constrained H2 optimization algorithm for mechatronics design in flexure-linked biaxial gantry.

    Science.gov (United States)

    Ma, Jun; Chen, Si-Lu; Kamaldin, Nazir; Teo, Chek Sing; Tay, Arthur; Mamun, Abdullah Al; Tan, Kok Kiong

    2017-11-01

    The biaxial gantry is widely used in many industrial processes that require high precision Cartesian motion. The conventional rigid-link version suffers from breaking down of joints if any de-synchronization between the two carriages occurs. To prevent above potential risk, a flexure-linked biaxial gantry is designed to allow a small rotation angle of the cross-arm. Nevertheless, the chattering of control signals and inappropriate design of the flexure joint will possibly induce resonant modes of the end-effector. Thus, in this work, the design requirements in terms of tracking accuracy, biaxial synchronization, and resonant mode suppression are achieved by integrated optimization of the stiffness of flexures and PID controller parameters for a class of point-to-point reference trajectories with same dynamics but different steps. From here, an H 2 optimization problem with defined constraints is formulated, and an efficient iterative solver is proposed by hybridizing direct computation of constrained projection gradient and line search of optimal step. Comparative experimental results obtained on the testbed are presented to verify the effectiveness of the proposed method. Copyright © 2017 ISA. Published by Elsevier Ltd. All rights reserved.

  4. TU-CD-304-04: Scanning Field Total Body Irradiation Using Dynamic Arc with Variable Dose Rate and Gantry Speed

    Energy Technology Data Exchange (ETDEWEB)

    Yi, B; Xu, H; Mutaf, Y; Prado, K [Univ. of Maryland School Of Medicine, Baltimore, MD (United States)

    2015-06-15

    Purpose: Enable a scanning field total body irradiation (TBI) technique, using dynamic arcs, which is biologically equivalent to a moving couch TBI. Methods: Patient is treated slightly above the floor and the treatment field scans across the patient by a moving gantry. MLC positions change during gantry motion to keep same field opening at the level of the treatment plane (170 cm). This is done to mimic the same geometry as the moving couch TBI technique which has been used in our institution for over 10 years. The dose rate and the gantry speed are determined considering a constant speed of the moving field, variations in SSD and slanted depths resulting from oblique gantry angles. An Eclipse (Varian) planning system is commissioned to accommodate the extended SSD. The dosimetric foundations of the technique have been thoroughly investigated using phantom measurements. Results: Dose uniformity better than 2% across 180 cm length at 10cm depth is achieved by moving the gantry from −55 to +55 deg. Treatment range can be extended by increasing gantry range. No device such as a gravity-oriented compensator is needed to achieve a uniform dose. It is feasible to modify the dose distribution by adjusting the dose rate at each gantry angle to compensate for body thickness differences. Total treatment time for 2 Gy AP/PA fields is 40–50 minutes excluding patient set up time, at the machine dose rate of 100 MU/min. Conclusion: This novel yet transportable moving field technique enables TBI treatment in a small treatment room with less program development preparation than other techniques. Treatment length can be extended per need, and. MLC-based thickness compensation and partial lung blocking are also possible.

  5. Design of a non-linear power take-off simulator for model testing of rotating wave energy devices

    Energy Technology Data Exchange (ETDEWEB)

    Lopes, M.F.P.; Henriques, J.C.C.; Lopes, Miguel C.; Gato, L.M.C. [IDMEC - Instituto de Engenharia Mecanica, Instituto Superior Tecnico, Universidade Tecnica de Lisboa, Lisboa (Portugal); Dente Antonio [CIE3 - Center for Innovation in Electrical and Energy Engineering, Lisboa (Portugal)

    2009-07-01

    Eddy current brakes provide a versatile way of simulating the power take-off system (PTO) in the model testing of wave energy converters at small scale. These are based on the principle that a conductive material moving perpendicularly to a magnetic field generates a braking force proportional to its velocity. This was applied in the design of the PTO simulator of a bottom-hinged flap wave energy converter model, at 1/16 scale. The efforts put into the accurate dynamic simulation of the device led to the development of a controllable PTO simulator, which can be applied to other small scale rotating wave energy device models. A special power source was built to provide the required controllable current intensity to feed the magnetic field generating coils. Different non-linear damping PTO characteristic curves can be simulated by basing the current control on real-time velocity measurement. The calibration of the system was done by connecting the device to a constant rotating speed motor and measuring the resistent torque produced by the PTO with a torquemeter for different values of current intensity through the coils.

  6. Markerless PET motion correction: tracking in narrow gantries through optical fibers

    DEFF Research Database (Denmark)

    Jensen, Rasmus Ramsbøl; Olesen, Oline Vinter; Benjaminsen, Claus

    2015-01-01

    be accurate while only adding minimal complexity to the workflow. We present: Tracoline 2.0, a surface scanner prototype, which allows for markerless tracking in the clinic. The system uses structured light through optical fibre bundles, which easily fit in narrow gantries. The optical fibres also makes...... the system compatible with magnetic resonance (MR) imaging since all the electronics are moved away from the scanner. We demonstrate the system in a positron emission tomography (PET) study using the Siemens high resolution research tomography (HRRT). With two Ge/Ga-68 line sources fitted in a mannequin head...... for rotations up to ±25º. Based on the tracking results the PET frames were also successfully corrected for motion by aligning 10 s frames without motion for the stepwise experiment and aligning 1 s frames for the experiment with continuous motion. We have demonstrated and evaluated a system for markerless...

  7. A method, device and application for the dynamic balancing of a rotating component

    International Nuclear Information System (INIS)

    Voinis, P.

    1995-01-01

    The dynamic balancing method is based on the detection of the vibrations generated by an unbalance; two satellites are then displaced in order to create a counter-unbalance and their position is measured. Their position is then adjusted so as the unbalance and counter-unbalance phases and intensities differences are inferior to predetermined reference values in order to balance dynamically the rotating component. Application to superpower turbogenerator shafting systems. 4 fig

  8. Optimizing 4D cone beam computed tomography acquisition by varying the gantry velocity and projection time interval

    International Nuclear Information System (INIS)

    O’Brien, Ricky T; Cooper, Benjamin J; Keall, Paul J

    2013-01-01

    Four dimensional cone beam computed tomography (4DCBCT) is an emerging clinical image guidance strategy for tumour sites affected by respiratory motion. In current generation 4DCBCT techniques, both the gantry rotation speed and imaging frequency are constant and independent of the patient’s breathing which can lead to projection clustering. We present a mixed integer quadratic programming (MIQP) model for respiratory motion guided-4DCBCT (RMG-4DCBCT) which regulates the gantry velocity and projection time interval, in response to the patient’s respiratory signal, so that a full set of evenly spaced projections can be taken in a number of phase, or displacement, bins during the respiratory cycle. In each respiratory bin, an image can be reconstructed from the projections to give a 4D view of the patient’s anatomy so that the motion of the lungs, and tumour, can be observed during the breathing cycle. A solution to the full MIQP model in a practical amount of time, 10 s, is not possible with the leading commercial MIQP solvers, so a heuristic method is presented. Using parameter settings typically used on current generation 4DCBCT systems (4 min image acquisition, 1200 projections, 10 respiratory bins) and a sinusoidal breathing trace with a 4 s period, we show that the root mean square (RMS) of the angular separation between projections with displacement binning is 2.7° using existing constant gantry speed systems and 0.6° using RMG-4DCBCT. For phase based binning the RMS is 2.7° using constant gantry speed systems and 2.5° using RMG-4DCBCT. The optimization algorithm presented is a critical step on the path to developing a system for RMG-4DCBCT. (paper)

  9. Motion control of a gantry crane with a container

    Science.gov (United States)

    Shugailo, T. S.; Yushkov, M. P.

    2018-05-01

    The transportation of a container by a gantry crane in a given time from one point of space to another is considered. The system is at rest at the end of the motion. A maximum admissible speed is taken into account. The control force is found using either the Pontryagin maximum principle or the generalized Gauss principle. The advantages of the second method over the first one is demonstrated.

  10. The development of a mobile CT-scanner gantry for use in the operating room

    International Nuclear Information System (INIS)

    Okudera, Hiroshi; Kobayashi, Shigeaki; Koike, Jouji; Harada, Takanobu; Kanemaru, Kei

    1989-01-01

    We report the development of a mobile CT-scanner gantry which uses a gantry platter. This system has been developed for use in the operating room. We designed a small lift to move the gantry unit of the scanner: the gantry carrier. The scanner gantry is fixed to the gantry carrier. A phantom test with a digitalized operating table worked well in the laboratory, and operating-room use showed that there was no deterioration in image quality. The mobile gantry system has been developed to increase the efficiency of the operating CT-scanner system. This system enables us to obtain CT images during surgery of immediately after surgery in the operating room, i.e., in cases that are not transferable to the radiological department. The operability is basically the same as that of a conventional mobile X-ray unit. Theoretically, this unit could be used with any CT scanner and in any operating room. (author)

  11. Fuzzy Controllers for a Gantry Crane System with Experimental Verifications

    Directory of Open Access Journals (Sweden)

    Naif B. Almutairi

    2016-01-01

    Full Text Available The control problem of gantry cranes has attracted the attention of many researchers because of the various applications of these cranes in the industry. In this paper we propose two fuzzy controllers to control the position of the cart of a gantry crane while suppressing the swing angle of the payload. Firstly, we propose a dual PD fuzzy controller where the parameters of each PD controller change as the cart moves toward its desired position, while maintaining a small swing angle of the payload. This controller uses two fuzzy subsystems. Then, we propose a fuzzy controller which is based on heuristics. The rules of this controller are obtained taking into account the knowledge of an experienced crane operator. This controller is unique in that it uses only one fuzzy system to achieve the control objective. The validity of the designed controllers is tested through extensive MATLAB simulations as well as experimental results on a laboratory gantry crane apparatus. The simulation results as well as the experimental results indicate that the proposed fuzzy controllers work well. Moreover, the simulation and the experimental results demonstrate the robustness of the proposed control schemes against output disturbances as well as against uncertainty in some of the parameters of the crane.

  12. Gantry Crane Scheduling with Interference Constraints in Railway Container Terminals

    Directory of Open Access Journals (Sweden)

    Peng Guo

    2013-04-01

    Full Text Available Railway container terminals, where gantry cranes are responsible for loading and unloading containers between freight trains and yards, are important hubs of hinterland logistics transportation. Terminal managers confront the challenge in improving the efficiency of their service. As the most expensive equipment in a terminal, the operational performance of gantry cranes is a crucial factor. In this paper, the gantry crane scheduling problem of railway container terminals is investigated. A mixed integer programming model which considers the effect of dwelling position dependent processing times is formulated. In addition, the safety distances, the travel times and the non-crossing requirement of cranes are incorporated in the mathematical model. A novel discrete artificial bee colony algorithm is presented to solve the intractable scheduling problem. Computational experiments are conducted to evaluate the proposed algorithm on some randomly constructed instances based on typical terminal operational data. Experimental results show that the proposed approach can obtain near optimal solutions for the investigated problem in a reasonable computational time.

  13. [Efficiency of a postoperative treatment after rotator cuff repair with a continuous passive motion device (CPM)].

    Science.gov (United States)

    Michael, J W-P; König, D P; Imhoff, A B; Martinek, V; Braun, S; Hübscher, M; Koch, C; Dreithaler, B; Bernholt, J; Preis, S; Loew, M; Rickert, M; Speck, M; Bös, L; Bidner, A; Eysel, P

    2005-01-01

    The main objective of this study was to prove that a postoperative combined continuous passive motion (CPM) and physiotherapy treatment protocol (CPM group) can achieve 90 degrees active abduction in the shoulder joint earlier than physiotherapy alone (PT group). The indication was a complete tear of the rotator cuff. The study was conducted under in-patient and out-patient conditions. 55 patients were included in this study. The prospective, randomized multicenter study design complies with DIN EN 540. The primary endpoint was the time span until 90 degrees active abduction was achieved by the patients. Patients in the CPM group reached the primary endpoint on average 12 days earlier than the control group. This difference was statistically significant (p = 0.0292). Analyzing the secondary endpoints, e. g., pain and disablement, the results in the CPM group showed again advantages of the combined treatment protocol (CPM + physiotherapy). The postoperative treatment of a total tear of the rotator cuff with a combined continuous passive motion and physiotherapy protocol provided a significantly earlier range of motion in the shoulder joint than physiotherapy alone. There was no report of CPM-related adverse effects.

  14. Security Analysis of Measurement-Device-Independent Quantum Key Distribution in Collective-Rotation Noisy Environment

    Science.gov (United States)

    Li, Na; Zhang, Yu; Wen, Shuang; Li, Lei-lei; Li, Jian

    2018-01-01

    Noise is a problem that communication channels cannot avoid. It is, thus, beneficial to analyze the security of MDI-QKD in noisy environment. An analysis model for collective-rotation noise is introduced, and the information theory methods are used to analyze the security of the protocol. The maximum amount of information that Eve can eavesdrop is 50%, and the eavesdropping can always be detected if the noise level ɛ ≤ 0.68. Therefore, MDI-QKD protocol is secure as quantum key distribution protocol. The maximum probability that the relay outputs successful results is 16% when existing eavesdropping. Moreover, the probability that the relay outputs successful results when existing eavesdropping is higher than the situation without eavesdropping. The paper validates that MDI-QKD protocol has better robustness.

  15. Optimum design of 6-DOF parallel manipulator with translational/rotational workspaces for haptic device application

    Energy Technology Data Exchange (ETDEWEB)

    Yoon, Jung Won; Hwang, Yoon Kwon [Gyeongsang National University, Jinju (Korea, Republic of); Ryu, Je Ha [Gwangju Institute of Science and Technology, Gwangju (Korea, Republic of)

    2010-05-15

    This paper proposes an optimum design method that satisfies the desired orientation workspace at the boundary of the translation workspace while maximizing the mechanism isotropy for parallel manipulators. A simple genetic algorithm is used to obtain the optimal linkage parameters of a six-degree-of-freedom parallel manipulator that can be used as a haptic device. The objective function is composed of a desired spherical shape translation workspace and a desired orientation workspace located on the boundaries of the desired translation workspace, along with a global conditioning index based on a homogeneous Jacobian matrix. The objective function was optimized to satisfy the desired orientation workspace at the boundary positions as translated from a neutral position of the increased entropy mechanism. An optimization result with desired translation and orientation workspaces for a haptic device was obtained to show the effectiveness of the suggested scheme, and the kinematic performances of the proposed model were compared with those of a preexisting base model

  16. Optimum design of 6-DOF parallel manipulator with translational/rotational workspaces for haptic device application

    International Nuclear Information System (INIS)

    Yoon, Jung Won; Hwang, Yoon Kwon; Ryu, Je Ha

    2010-01-01

    This paper proposes an optimum design method that satisfies the desired orientation workspace at the boundary of the translation workspace while maximizing the mechanism isotropy for parallel manipulators. A simple genetic algorithm is used to obtain the optimal linkage parameters of a six-degree-of-freedom parallel manipulator that can be used as a haptic device. The objective function is composed of a desired spherical shape translation workspace and a desired orientation workspace located on the boundaries of the desired translation workspace, along with a global conditioning index based on a homogeneous Jacobian matrix. The objective function was optimized to satisfy the desired orientation workspace at the boundary positions as translated from a neutral position of the increased entropy mechanism. An optimization result with desired translation and orientation workspaces for a haptic device was obtained to show the effectiveness of the suggested scheme, and the kinematic performances of the proposed model were compared with those of a preexisting base model

  17. Dynamic analysis of the gantry crane used for transporting BOP

    Directory of Open Access Journals (Sweden)

    Stanisław WOJCIECH

    2008-01-01

    Full Text Available In the paper the dynamic analysis of a gantry crane used for transporting of BOP (BlowOut Preventer is presented. The crane is placed on a drilling platform. Sea waves cause motion of the platform and the load. Description of such systems can be used in the design process of control systems which allows us to compensate waving. Homogenous transformations and joint coordinates are used to describe behavior of the system. Equations of motion are derived using the Lagrange equations of the second order. In the paper the results of numerical calculations are presented as well.

  18. Conceptual and ion-optical designs of an isocentric gantry for light-ion cancer therapy

    International Nuclear Information System (INIS)

    Vorobiev, L.G.; Weick, H.; Wollnik, H.

    1998-02-01

    Conceptual and ion-optical designs of an isocentric gantry are presented. The gantry is designed for light-ion therapy using a 6 Tm carbon beam. Its main characteristics are: 5 m overall radius, 1.4 m drift to the patient, 20x10 cm 2 treatment field, and two-direction active parallel scanning of a fine focused pencil beam. To achieve these features, the beam scanners are optimally positioned upstream the last 90 bending section consisting of two 45 sector magnets with 4 oblique field boundaries and different apertures. The gantry concept is a compromise between 'barrel' and 'conical' gantry shapes and is likely the most space-saving configuration of single-plane isocentric gantries. This new so-called 'short barrel' configuration features also small aperture quadruple lenses, low intermediate dispersion and a fully achromatic beam transport system. Its overall dimensions stay close to the size of the existing proton gantries. (orig.)

  19. A novel beam optics concept in a particle therapy gantry utilizing the advantages of superconducting magnets.

    Science.gov (United States)

    Gerbershagen, Alexander; Meer, David; Schippers, Jacobus Maarten; Seidel, Mike

    2016-09-01

    A first order design of the beam optics of a superconducting proton therapy gantry beam is presented. The possibilities of superconducting magnets with respect to the beam optics such as strong fields, large apertures and superposition of different multipole fields have been exploited for novel concepts in a gantry. Since various techniques used in existing gantries have been used in our first design steps, some examples of the existing superconducting gantry designs are described and the necessary requirements of such a gantry are explained. The study of a gantry beam optics design is based on superconducting combined function magnets. The simulations have been performed in first order with the conventional beam transport codes. The superposition of strong dipole and quadrupole fields generated by superconducting magnets enables the introduction of locally achromatic bending sections without increasing the gantry size. A rigorous implementation of such beam optics concepts into the proposed gantry design dramatically increases the momentum acceptance compared to gantries with normal conducting magnets. In our design this large acceptance has been exploited by the implementation of a degrader within the gantry and a potential possibility to use the same magnetic field for all energies used in a treatment, so that the superconducting magnets do not have to vary their fields during a treatment. This also enables very fast beam energy changes, which is beneficial for spreading the Bragg peak over the thickness of the tumor. The results show an improvement of its momentum acceptance. Large momentum acceptance in the gantry creates a possibility to implement faster dose application techniques. Copyright © 2016. Published by Elsevier GmbH.

  20. Beam-optics study of the gantry beam delivery system for light-ion cancer therapy

    International Nuclear Information System (INIS)

    Pavlovic, M.

    1995-12-01

    Ion optics considerations on the granty-like beam delivery system for light-ion cancer therapy are presented. A low-angle active beam scanning in two directions is included in the preliminary gantry design. The optical properties of several gantry modifications are discussed. (orig.)

  1. Modular design of a scanning Gantry for animal SPECT, PET and CT applications

    International Nuclear Information System (INIS)

    Brady, J.A.M.; Lerch, M.; Bourke, J.; Rozenfeld, A.B.; Meikle, S.R.

    2005-01-01

    Full text: Using the modular design methodology, we have developed a scanning gantry system; custom designed for multi-modality nuclear imaging techniques (NITs) and the CoALA SPECT project. This paper discusses the gantry's flexible modular design and how significant components and functionality have been reused in other medical physics related instrumentation developed at the Centre for Medical Radiation Physics, University of Wollongong. The gantry is designed to meet the exacting requirements of researchers in new radiotracer development who require accurate functional and anatomical information-the scanning gantry has a linear stepping accuracy of +/-12.5 microns and an angular accuracy of +/-0.03 degree. The scanning gantry is a precision instrument that is also affordable to research groups working in small laboratories or universities. Copyright (2005) Australian Institute of Physics

  2. 78 FR 68477 - Overhead and Gantry Cranes; Extension of the Office of Management and Budget's (OMB) Approval of...

    Science.gov (United States)

    2013-11-14

    ...] Overhead and Gantry Cranes; Extension of the Office of Management and Budget's (OMB) Approval of... requirements specified in the Standard on Overhead and Gantry Cranes (29 CFR 1910.179). DATES: Comments must be... information (29 U.S.C. 657). The paperwork provisions of the Standard on Overhead and Gantry Cranes specify...

  3. A light-weight compact proton gantry design with a novel dose delivery system for broad-energetic laser-accelerated beams.

    Science.gov (United States)

    Masood, U; Cowan, T E; Enghardt, W; Hofmann, K M; Karsch, L; Kroll, F; Schramm, U; Wilkens, J J; Pawelke, J

    2017-07-07

    parameters, deliverable via the presented gantry and ELPIS dose delivery system. The conventional PT gantries are huge and require large space for the gantry to rotate the beam around the patient, which could be reduced up to 4 times with the presented pulse powered gantry system. The further developments in the next generation petawatt laser systems and laser-targets are crucial to reach higher proton energies. However, if proton energies required for therapy applications are reached it could be possible in future to reduce the footprint of the PT facilities, without compromising on clinical standards.

  4. A light-weight compact proton gantry design with a novel dose delivery system for broad-energetic laser-accelerated beams

    Science.gov (United States)

    Masood, U.; Cowan, T. E.; Enghardt, W.; Hofmann, K. M.; Karsch, L.; Kroll, F.; Schramm, U.; Wilkens, J. J.; Pawelke, J.

    2017-07-01

    parameters, deliverable via the presented gantry and ELPIS dose delivery system. The conventional PT gantries are huge and require large space for the gantry to rotate the beam around the patient, which could be reduced up to 4 times with the presented pulse powered gantry system. The further developments in the next generation petawatt laser systems and laser-targets are crucial to reach higher proton energies. However, if proton energies required for therapy applications are reached it could be possible in future to reduce the footprint of the PT facilities, without compromising on clinical standards.

  5. A quality assurance device for the accuracy of the isocentres of teletherapy and simulation machines

    International Nuclear Information System (INIS)

    Arjomandy, Bijan; Altschuler, Martin D.

    2000-01-01

    A new quality assurance device has been designed to measure the location and wobble of the radiation isocentre of linacs and simulation machines as a function of gantry rotation. The radiation isocentre is the intersection in space of the central x-rays of a linac or simulation machine at different gantry angles. Six radio-opaque markers 1 mm in size are embedded in a radio-transparent calibration object (specifically, a hollowed cube) in such a way that the markers are non-coplanar and uniquely identifiable in radiographic projections. The projective radiographs are obtained on the films held by the film holder attached to the gantry during the QA procedure. The marker positions of the calibration object define a known 3D reference frame, and their image positions on each radiograph determine the projective (3D to 2D matrix) transformation for that radiograph. Once the transformation is found, a 3D ray from the radiation source to any radiograph pixel becomes known. The radiographic pixels are coordinated (positioned and scaled) with respect to the projected image of radio-opaque fiducial cross-hairs fixed to a block tray and thus to the gantry. We select the central ray to correspond to the radiographic pixel whose rays at different gantry angles intersect in the smallest spatial domain. That pixel is found by a spiral search in the radiograph outward from the image of the radio-opaque cross-hair intersection. The wobble of the isocentre is defined by the set of points (on the central rays) at closest approach to the isocentre. The device was tested and compared with commercially available QA devices. It is able to locate the isocentre to within 0.5 mm. The offset of this derived radiation isocentre from the intersection of the positioning lasers can be found. To do this, the calibration object is initially placed so that the laser intersection point falls on a seventh radio-opaque marker near the centre of the hollow cube calibration object. The seventh marker

  6. Operator control systems and methods for swing-free gantry-style cranes

    Science.gov (United States)

    Feddema, John T.; Petterson, Ben J.; Robinett, III, Rush D.

    1998-01-01

    A system and method for eliminating swing motions in gantry-style cranes while subject to operator control is presented. The present invention comprises an infinite impulse response ("IIR") filter and a proportional-integral ("PI") feedback controller (50). The IIR filter receives input signals (46) (commanded velocity or acceleration) from an operator input device (45) and transforms them into output signals (47) in such a fashion that the resulting motion is swing free (i.e., end-point swinging prevented). The parameters of the IIR filter are updated in real time using measurements from a hoist cable length encoder (25). The PI feedback controller compensates for modeling errors and external disturbances, such as wind or perturbations caused by collision with objects. The PI feedback controller operates on cable swing angle measurements provided by a cable angle sensor (27). The present invention adjusts acceleration and deceleration to eliminate oscillations. An especially important feature of the present invention is that it compensates for variable-length cable motions from multiple cables attached to a suspended payload.

  7. Real-Time Dispatching of Rubber Tired Gantry Cranes in Container Terminals

    National Research Council Canada - National Science Library

    McNary, Bradley S

    2008-01-01

    ... significant growth in container traffic. Efficiently managing of rubber tired gantry cranes and planning container placement within the terminal are two ways to increase the overall efficiency of a terminal...

  8. Influence of gantry angle in helical computed tomography. Usefullness of 1-dimension sharpness filter

    Energy Technology Data Exchange (ETDEWEB)

    Kawano, Hirofumi; Suzuki, Toru; Matsuura, Shigeru; Kai, Tsuyoshi; Shimizu, Toshiyuki [Hyuga Hospital of Saiseikai Foundation, Kadogawa, Miyazaki (Japan)

    2001-05-01

    When we let gantry tilt and do scan in helical CT, vertical sharpness deteriorates. We were able to revise it with 1-dimensional sharpness filter which the square sum of difference of MTF was compared, and was designed this time. And the unsharpness was in proportion to sin of gantry angle. As a result, we led several sets of frequency emphasis degree. There is a model to built 1-dimension sharpness filter in a scan plan. It is useful for clinical diagnoses. (author)

  9. Influence of gantry angle in helical computed tomography. Usefullness of 1-dimension sharpness filter

    International Nuclear Information System (INIS)

    Kawano, Hirofumi; Suzuki, Toru; Matsuura, Shigeru; Kai, Tsuyoshi; Shimizu, Toshiyuki

    2001-01-01

    When we let gantry tilt and do scan in helical CT, vertical sharpness deteriorates. We were able to revise it with 1-dimensional sharpness filter which the square sum of difference of MTF was compared, and was designed this time. And the unsharpness was in proportion to sin of gantry angle. As a result, we led several sets of frequency emphasis degree. There is a model to built 1-dimension sharpness filter in a scan plan. It is useful for clinical diagnoses. (author)

  10. Mechanical characterization of the Varian Exact-arm and R-arm support systems for eight aS500 electronic portal imaging devices

    International Nuclear Information System (INIS)

    Grattan, Mark W. D.; McGarry, Conor K.

    2010-01-01

    Purpose: The aim of this study is to compare the positioning accuracy at different gantry angles of two electronic portal imaging devices (EPIDs) support arm systems by using EPID difference images as a measure for displacement. This work presents a comparison of the mechanical performance of eight Varian aS500 (Varian Medical Systems, Palo Alto, CA) EPIDs, mounted using either the Varian Exact-arm or R-arm. Methods: The mechanical performance of the two arm systems was compared by investigating the variation in sensitivity with gantry angle, both before and after the EPID position was adjusted after gantry rotation. Positional errors were investigated by subtracting images from a reference image taken at gantry 0 deg., and the amplitude of the peaks and troughs at the field edges for longitudinal (radial) and lateral (transverse) profiles across the resulting image was related to the distance of displacement. Calibration curves based on a pixel-by-pixel shift were generated for each EPID and the Varian hand pendant accuracy was compared to the calibration data. Results: The response of the EPIDs was found to change with gantry rotation, with the largest difference at 180 deg. The Exact-arm was found to correct well for any displacement, while the R-arm tended to overcorrect following repositioning using the hand pendant. The calibration curves were consistent within each set of matched linacs, and the hand pendant accuracy was similar for both arm systems, although generally in different directions. With respect to gantry rotation effects, the mechanical performance of the Exact-arm systems was found to be much better than that of the R-arm systems. At gantry positions 90 deg., 270 deg., and 180 deg. the average misalignment in the longitudinal direction was +4.2±0.2, +1.8±1.6, and +7.4±0.5 mm for the R-arms, and +2.9±0.2, +2.1±0.8, and +4.9±0.7 mm for the Exact-arms. In the lateral direction the average positional errors were +2.1±0.4, -4.7±0.4, and -2.5

  11. A novel gantry for proton therapy at the Paul Scherrer Institute

    International Nuclear Information System (INIS)

    Pedroni, E.; Boehringer, T.; Coray, A.; Goitein, G.; Grossmann, M.; Lomax, A.; Lin, S.; Jermann, M.

    2001-01-01

    PSI has gained in the last few years the unique experience of using a proton therapy system based on a beam scanning delivery technique and on a compact gantry. This knowledge is now bringing forth new initiatives. We are continuously producing significant modifications and improvements to the present system, gantry 1. The major new step is however the decision of PSI to purchase a dedicated accelerator for the medical project. In the context of the expansion of the medical project of PSI (project PROSCAN) we have also started to plan the realisation of a second proton gantry, gantry 2. In this lecture we present the main ideas for the novel gantry, which will be based on one hand on the experience with the present technology, but on the other hand should be designed as a system more open to further developments and needs. The established and the future requirements for the beam delivery on the new gantry were routed into the specification list for the dedicated accelerator

  12. Preliminary results for positron emission mammography: real-time functional breast imaging in a conventional mammography gantry

    International Nuclear Information System (INIS)

    Weinberg, I.; Majewski, S.; Weisenberger, A.; Markowitz, A.; Aloj, L.; Majewski, L,; Danforth, D.; Mulshine, J.; Cowan, K.; Zujewski, J.; Chow, C.; Jones, E.; Chang, V.; Berg, W.; Frank, J.

    1996-01-01

    In order to optimally integrate radiotracer breast imaging within the breast clinic, anatomy and pathology should be easily correlated with functional nuclear medicine breast images. As a first step in the development of a hybrid functional/anatomic breast imaging platform with biopsy capability, a conventional X-ray mammography gantry was modified to image the compressed brest with positron emitters. Phantom studies with the positron emission mammography (PEM) device showed that a 1-cc hot spot could be detected within 5 min. A preliminary clinical trial demonstrated in vivo visualization of primary breast cancer within 4 min. For sites where positron-emitting radionuclides are available, PEM promises to achieve low-cost directed functional examination of breast abnormalities, with potential for achieving X-ray correlation and image-guided biopsy. (orig.)

  13. Rotating preventers

    International Nuclear Information System (INIS)

    Tangedahl, M.J.; Stone, C.R.

    1992-01-01

    This paper reports that recent changes in the oil and gas industry and ongoing developments in horizontal and underbalanced drilling necessitated development of a better rotating head. A new device called the rotating blowout preventer (RBOP) was developed by Seal-Tech. It is designed to replace the conventional rotating control head on top of BOP stacks and allows drilling operations to continue even on live (underbalanced) wells. Its low wear characteristics and high working pressure (1,500 psi) allow drilling rig crews to drill safely in slightly underbalanced conditions or handle severe well control problems during the time required to actuate other BOPs in the stack. Drilling with a RBOP allows wellbores to be completely closed in tat the drill floor rather than open as with conventional BOPs

  14. Effects of Interfaces on Dynamics in Micro-Fluidic Devices: Slip-Boundaries’ Impact on Rotation Characteristics of Polar Liquid Film Motors

    Science.gov (United States)

    Jiang, Su-Rong; Liu, Zhong-Qiang; Amos Yinnon, Tamar; Kong, Xiang-Mu

    2017-05-01

    A new approach for exploring effects of interfaces on polar liquids is presented. Their impact on the polar liquid film motor (PLFM) - a novel micro-fluidic device - is studied. We account for the interface’s impact by modeling slip boundary effects on the PLFM’s electro-hydro-dynamical rotations. Our analytical results show as k={l}s/R increases (with {l}s denoting the slip length resulting from the interface’s impact on the film’s properties, k > -1 and R denoting the film’s radius): (a) PLFMs subsequently exhibit rotation characteristics under “negative-”, “no-”, “partial-” and “perfect-” slip boundary conditions; (b) The maximum value of the linear velocity of the steady rotating film increases linearly and its location approaches the film’s border; (c) The decay of the angular velocities’ dependency on the distance from the center of the film slows down, resulting in a macroscopic flow near the boundary. With our calculated rotation speed distributions consistent with the existing experimental ones, research aiming at fitting computed to measured distributions promises identifying the factors affecting {l}s, e.g., solid-fluid potential interactions and surface roughness. The consistency also is advantageous for optimizing PLFM’s applications as micro-washers, centrifuges, mixers in the lab-on-a-chip. Supported by National Natural Science Foundation of China under Grant Nos. 11302118, 11275112, and Natural Science Foundation of Shandong Province under Grant No. ZR2013AQ015

  15. Control Strategy for Automatic Gantry Crane Systems: A Practical and Intelligent Approach

    Directory of Open Access Journals (Sweden)

    Wahyudi

    2007-12-01

    Full Text Available The use of gantry crane systems for transporting payload is very common in building constructions. However, moving the payload using the crane is not an easy task especially when strict specifications on the swing angle and on the transfer time need to be satisfied. Various attempts in controlling gantry cranes system based on open- loop and closed-loop control systems were proposed. However, most of the proposed controllers were designed based on the model and parameter of the crane system. In general, modeling and parameter identifications are troublesome and time consuming task. To overcome this problem, in this paper, a practical and intelligent control method for automatic gantry crane is introduced and evaluated experimentally. The results show that the proposed method is not only effective for controlling the crane but also robust to parameter variation.

  16. Control Strategy for Automatic Gantry Crane Systems: A Practical and Intelligent Approach

    Directory of Open Access Journals (Sweden)

    Wahyudi

    2008-11-01

    Full Text Available The use of gantry crane systems for transporting payload is very common in building constructions. However, moving the payload using the crane is not an easy task especially when strict specifications on the swing angle and on the transfer time need to be satisfied. Various attempts in controlling gantry cranes system based on open- loop and closed-loop control systems were proposed. However, most of the proposed controllers were designed based on the model and parameter of the crane system. In general, modeling and parameter identifications are troublesome and time consuming task. To overcome this problem, in this paper, a practical and intelligent control method for automatic gantry crane is introduced and evaluated experimentally. The results show that the proposed method is not only effective for controlling the crane but also robust to parameter variation.

  17. X-band Linac for a 6 MeV dual-head radiation therapy gantry

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Seung Hyun; Shin, Seung-Wook; Lee, Jongchul; Kim, Hui-Su [WCU Department of Energy Science, Suwon 440-746 (Korea, Republic of); Lee, Byeong-No; Lee, Byung-Chul [Radiation Instrumentation Research Division, Korea Atomic Energy Research Institute, Jeongeup 56212 (Korea, Republic of); Park, Hyung-dal; Song, Ki-back [Radiation Technology eXcellence (RTX), Daejeon 305-500 (Korea, Republic of); Song, Ho-seung; Mun, Sangchul; Ha, Donghyup [School of Information and Communication Engineering, Sungkyunkwan University, Suwon 440-746 (Korea, Republic of); Chai, Jong-Seo, E-mail: jschai@skku.edu [School of Information and Communication Engineering, Sungkyunkwan University, Suwon 440-746 (Korea, Republic of)

    2017-04-21

    We developed a design for a 6 MeV X-band linear accelerator for radiation therapy in a dual-head gantry layout. The dual-head gantry has two linacs that can be operated independently. Each X-band linac accelerates electron bunches using high-power RF and generates X-rays for radiation therapy. It requires a versatile RF system and pulse sequence to accomplish various radiation therapy procedures. The RF system consists of 9.3 GHz, 2 MW X-band magnetron and associated RF transmission components. A test linac was assembled and operated to characterize its RF performance without beam. This paper presents these results along with a description of the gantry linacs and their operational requirements.

  18. An improved method to accurately calibrate the gantry angle indicators of the radiotherapy linear accelerators

    International Nuclear Information System (INIS)

    Chang Liyun; Ho, S.-Y.; Du, Y.-C.; Lin, C.-M.; Chen Tainsong

    2007-01-01

    The calibration of the gantry angle indicator is an important and basic quality assurance (QA) item for the radiotherapy linear accelerator. In this study, we propose a new and practical method, which uses only the digital level, V-film, and general solid phantoms. By taking the star shot only, we can accurately calculate the true gantry angle according to the geometry of the film setup. The results on our machine showed that the gantry angle was shifted by -0.11 deg. compared with the digital indicator, and the standard deviation was within 0.05 deg. This method can also be used for the simulator. In conclusion, this proposed method could be adopted as an annual QA item for mechanical QA of the accelerator

  19. Development of a permanently controllable rotating biopsy device. Pt. I. Theoretical considerations and in-vitro results for five different prototypes

    International Nuclear Information System (INIS)

    Schaefer, Philipp Jost; Jahnke, T.; Andres, H.; Heller, M.; Schaefer, F.K.; Hedderich, J.

    2009-01-01

    Purpose: to develop different prototypes of permanently controllable rotating biopsy devices with determination of the most efficient prototype in biopsies in bovine myocardium. Materials and Methods: Five different prototypes of 18-gauge rotating biopsy devices were designed and constructed, four (1-4) with various drill-like cutting edges and one (5) cannula type with a lancet-like helically bent cutting edge. Using bovine myocardium as the biopsy tissue, n = 100 specimens per prototype were obtained, and a quantitative analysis including tissue fragmentation, length in mm and weight in mg was carried out. For statistical analysis, the chi-square test for tissue fragmentation and Kruskal-Wallis test for the parameters length and weight were calculated. Results: prototype 5 showed the highest rate of extraction of one-fragment specimens in n=66 cases and the lowest rate of failure to obtain tissue in n=11 cases. The mean length/weight were 4.15 mm/3.91 mg for prototype 1, 1.80 mm/1.66 mg for prototype 2, 4.61 mm/3.28 mg for prototype 3, 5.20 mm/3.74 mg for prototype 4, and 9.57 mm/6.97 mg for prototype 5. In all three categories, prototype 5 was significantly superior to the prototypes 1-4 with p < 0.001. Conclusion: the cannula type with a lancet-like helically bent cutting edge proved to be the most efficient prototype and may now be tested competitively against established automated biopsy devices in vitro. (orig.)

  20. Supporting and guiding device that is leak-tight and can be dismantled for the shaft of a rotating machine

    International Nuclear Information System (INIS)

    Tigoulet, Bernard; Fanchtein, J.P.; Dubost, Rene.

    1982-01-01

    This device includes a removable bearing casing crossed by at least one shaft of the machine, facilities for guiding this casing in parallel with the axis of the shaft so as to facilitate its removal and refitting, a system for supporting the shaft when the removable casing is not fitted in the machine frame. Application to machines for the extrusion of coating bitumen for radioactive waste [fr

  1. 75 FR 44288 - Overhead and Gantry Cranes; Extension of the Office of Management and Budget's (OMB) Approval of...

    Science.gov (United States)

    2010-07-28

    ...] Overhead and Gantry Cranes; Extension of the Office of Management and Budget's (OMB) Approval of... requirements specified in the Standard on Overhead and Gantry Cranes (29 CFR 1910.179). DATES: Comments must be... requirements for: Marking the rated load of cranes; preparing certification records to verify the inspection of...

  2. SU-F-T-207: Does the Greater Flexibility of Pencil Beam Scanning Reduce the Need for a Proton Gantry?

    Energy Technology Data Exchange (ETDEWEB)

    Yan, S; Depauw, N; Flanz, J; Adams, J; Gorissen, BL; Shih, H; Bortfeld, T; Lu, H [Massachusetts General Hospital and Harvard Medical School, Department of Radiation Oncology, Boston, MA (United States)

    2016-06-15

    Purpose: Gantry-less proton treatment facility could lower the capital cost of proton therapy. This study investigates the dosimetric feasibility of using only coplanar pencil beam scanning (PBS) beams for those patients who had beam angles that would not have been deliverable without the gantry. Those coplanar beams are implemented on gantry-less horizontal beam-line with patients in sitting or standing positions. Methods: We have selected ten patients (seven head-and-neck, one thoracic, one abdominal and one pelvic case) with clinically delivered double scattering (DS) or PBS treatment plans with beam angles that were challenging to achieve without a gantry. After removing these beams angles, PBS plans were optimized for gantry-less intensity modulated proton therapy (IMPT) or single field optimization (SFO) with multi-criteria optimization (MCO). For head-and-neck patients who were treated by DS, we generated PBS plans with non-coplanar beams for comparison. Dose-volume-histograms (DVHs), target homogeneity index (HI), mean dose, D-2 and D-98 were reported. Robustness analysis was performed with ±2.5 mm setup errors and ±3.5% range uncertainties for three head-and-neck patients. Results: PBS-gantry-less plans provided more homogenous target coverage and significant improvements on organs-at-risk (OARs) sparing, compared to passive scattering treatments with a gantry. The PBS gantry-less treatments reduced the HI for target coverage by 1.3% to 47.2%, except for a suprasellar patient and a liver patient. The PBS-gantry-less plans reduced the D-mean of OARs by 3.6% to 67.4%. The PBS-gantry plans had similar target coverage and only marginal improvements on OAR sparing as compared to the PBS-gantry-less plans. These two PBS plans also had similar robustness relative to range uncertainties and setup errors. Conclusion: The gantry-less plans have with less mean dose to OARs and more homogeneous target coverage. Although the PBS-gantry plans have slightly improved

  3. Stabilizing model predictive control of a gantry crane based on flexible set-membership constraints

    NARCIS (Netherlands)

    Iles, Sandor; Lazar, M.; Kolonic, Fetah; Jadranko, Matusko

    2015-01-01

    This paper presents a stabilizing distributed model predictive control of a gantry crane taking into account the variation of cable length. The proposed algorithm is based on the off-line computation of a sequence of 1-step controllable sets and a condition that enables flexible convergence towards

  4. Alternating-gradient canted cosine theta superconducting magnets for future compact proton gantries

    Directory of Open Access Journals (Sweden)

    Weishi Wan

    2015-10-01

    Full Text Available We present a design of superconducting magnets, optimized for application in a gantry for proton therapy. We have introduced a new magnet design concept, called an alternating-gradient canted cosine theta (AG-CCT concept, which is compatible with an achromatic layout. This layout allows a large momentum acceptance. The 15 cm radius of the bore aperture enables the application of pencil beam scanning in front of the SC-magnet. The optical and dynamic performance of a gantry based on these magnets has been analyzed using the fields derived (via Biot-Savart law from the actual windings of the AG-CCT combined with the full equations of motion. The results show that with appropriate higher order correction, a large 3D volume can be rapidly scanned with little beam shape distortion. A very big advantage is that all this can be done while keeping the AG-CCT fields fixed. This reduces the need for fast field ramping of the superconducting magnets between the successive beam energies used for the scanning in depth and it is important for medical application since this reduces the technical risk (e.g., a quench associated with fast field changes in superconducting magnets. For proton gantries the corresponding superconducting magnet system holds promise of dramatic reduction in weight. For heavier ion gantries there may furthermore be a significant reduction in size.

  5. An Approximate Cone Beam Reconstruction Algorithm for Gantry-Tilted CT Using Tangential Filtering

    Directory of Open Access Journals (Sweden)

    Ming Yan

    2006-01-01

    Full Text Available FDK algorithm is a well-known 3D (three-dimensional approximate algorithm for CT (computed tomography image reconstruction and is also known to suffer from considerable artifacts when the scanning cone angle is large. Recently, it has been improved by performing the ramp filtering along the tangential direction of the X-ray source helix for dealing with the large cone angle problem. In this paper, we present an FDK-type approximate reconstruction algorithm for gantry-tilted CT imaging. The proposed method improves the image reconstruction by filtering the projection data along a proper direction which is determined by CT parameters and gantry-tilted angle. As a result, the proposed algorithm for gantry-tilted CT reconstruction can provide more scanning flexibilities in clinical CT scanning and is efficient in computation. The performance of the proposed algorithm is evaluated with turbell clock phantom and thorax phantom and compared with FDK algorithm and a popular 2D (two-dimensional approximate algorithm. The results show that the proposed algorithm can achieve better image quality for gantry-tilted CT image reconstruction.

  6. Anti-Swing Control of Gantry and Tower Cranes Using Fuzzy and Time-Delayed Feedback with Friction Compensation

    Directory of Open Access Journals (Sweden)

    H.M. Omar

    2005-01-01

    Full Text Available We designed a feedback controller to automate crane operations by controlling the load position and its swing. First, a PD tracking controller is designed to follow a prescribed trajectory. Then, another controller is added to the control loop to damp the load swing. The anti-swing controller is designed based on two techniques: a time-delayed feedback of the load swing angle and an anti-swing fuzzy logic controller (FLC. The rules of the FLC are generated by mapping the performance of the time-delayed feedback controller. The same mapping method used for generating the rules can be applied to mimic the performance of an expert operator. The control algorithms were designed for gantry cranes and then extended to tower cranes by considering the coupling between the translational and rotational motions. Experimental results show that the controller is effective in reducing load oscillations and transferring the load in a reasonable time. To experimentally validate the theory, we had to compensate for friction. To this end, we estimated the friction and then applied a control action to cancel it. The friction force was estimated by assuming a mathematical model and then estimating the model coefficients using an off-line identification technique, the method of least squares (LS.

  7. SU-E-J-56: Static Gantry Digital Tomosynthesis From the Beam’s-Eye-View

    International Nuclear Information System (INIS)

    Partain, L; Kwon, J; Boyd, D; Rottmann, J; Berbeco, R; Zentai, G

    2015-01-01

    Purpose We have designed a novel TumoTrak™ x-ray system that delivers 19 distinct kV views with the linac gantry stationary. It images MV treatment beam above and below the patient with a kV tomosysthesis slice image from the therapy beam’s-eye-view. Results will be high quality images without MLC shadowing for notable improvements relative to conventional fluoroscopic MV imaging and fluoroscopic kV imaging. Methods A complete design has a kV electron beam multisource X-ray tube that fits around the MV treatment beam path, with little interference with normal radiotherapy and unblocked by the multi-leaf-collimator. To simulate digital tomosynthesis, we used cone-beam CT projection data from a lung SBRT patient. These data were acquired at 125 kVp and 11 fps (0.4 mAs per projection). We chose 19 projections evenly spaced over 27° around one of the treatment angles (240°). Digital tomosynthesis reconstruction of a slice through the tumor was performed using iterative reconstruction. The visibility of the lesion was assessed for the reconstructed digital tomosynthesis (DTS), using fluoroscopy MV images acquired during radiation therapy, and a kV single projection image acquired at the same angle as the treatment field (240°). Results The fluoroscopic DTS images provide the best tumor contrast, surpassing the conventional radiographic and the in-treatment MV portal images. The electron beam multisource X-ray tube design has been completed and the tube is being fabricated. The estimated time to cycle through all 19 projections is 700 ms, enabling high frame-rate imaging. While the initial proposed use case is for image guided and gated treatment delivery, the enhanced imaging will also deliver superior radiographic images for patient setup. Conclusion The proposed device will deliver high quality planar images from the beam’s-eye-view without MLC obstruction. The prototype has been designed and is being assembled with first imaging scheduled for May 2015. L

  8. SU-E-J-56: Static Gantry Digital Tomosynthesis From the Beam’s-Eye-View

    Energy Technology Data Exchange (ETDEWEB)

    Partain, L; Kwon, J; Boyd, D [TeleSecurity Sciences, Las Vegas, NV (United States); Rottmann, J; Berbeco, R [Brigham and Woman’s Hospital, Boston, MA (United States); Zentai, G [Varian Medical Systems, Palo Alto, CA (United States)

    2015-06-15

    Purpose We have designed a novel TumoTrak™ x-ray system that delivers 19 distinct kV views with the linac gantry stationary. It images MV treatment beam above and below the patient with a kV tomosysthesis slice image from the therapy beam’s-eye-view. Results will be high quality images without MLC shadowing for notable improvements relative to conventional fluoroscopic MV imaging and fluoroscopic kV imaging. Methods A complete design has a kV electron beam multisource X-ray tube that fits around the MV treatment beam path, with little interference with normal radiotherapy and unblocked by the multi-leaf-collimator. To simulate digital tomosynthesis, we used cone-beam CT projection data from a lung SBRT patient. These data were acquired at 125 kVp and 11 fps (0.4 mAs per projection). We chose 19 projections evenly spaced over 27° around one of the treatment angles (240°). Digital tomosynthesis reconstruction of a slice through the tumor was performed using iterative reconstruction. The visibility of the lesion was assessed for the reconstructed digital tomosynthesis (DTS), using fluoroscopy MV images acquired during radiation therapy, and a kV single projection image acquired at the same angle as the treatment field (240°). Results The fluoroscopic DTS images provide the best tumor contrast, surpassing the conventional radiographic and the in-treatment MV portal images. The electron beam multisource X-ray tube design has been completed and the tube is being fabricated. The estimated time to cycle through all 19 projections is 700 ms, enabling high frame-rate imaging. While the initial proposed use case is for image guided and gated treatment delivery, the enhanced imaging will also deliver superior radiographic images for patient setup. Conclusion The proposed device will deliver high quality planar images from the beam’s-eye-view without MLC obstruction. The prototype has been designed and is being assembled with first imaging scheduled for May 2015. L

  9. Modelling and Metaheuristic for Gantry Crane Scheduling and Storage Space Allocation Problem in Railway Container Terminals

    Directory of Open Access Journals (Sweden)

    Ming Zeng

    2017-01-01

    Full Text Available The gantry crane scheduling and storage space allocation problem in the main containers yard of railway container terminal is studied. A mixed integer programming model which comprehensively considers the handling procedures, noncrossing constraints, the safety margin and traveling time of gantry cranes, and the storage modes in the main area is formulated. A metaheuristic named backtracking search algorithm (BSA is then improved to solve this intractable problem. A series of computational experiments are carried out to evaluate the performance of the proposed algorithm under some randomly generated cases based on the practical operation conditions. The results show that the proposed algorithm can gain the near-optimal solutions within a reasonable computation time.

  10. High precision tracking control of a servo gantry with dynamic friction compensation.

    Science.gov (United States)

    Zhang, Yangming; Yan, Peng; Zhang, Zhen

    2016-05-01

    This paper is concerned with the tracking control problem of a voice coil motor (VCM) actuated servo gantry system. By utilizing an adaptive control technique combined with a sliding mode approach, an adaptive sliding mode control (ASMC) law with friction compensation scheme is proposed in presence of both frictions and external disturbances. Based on the LuGre dynamic friction model, a dual-observer structure is used to estimate the unmeasurable friction state, and an adaptive control law is synthesized to effectively handle the unknown friction model parameters as well as the bound of the disturbances. Moreover, the proposed control law is also implemented on a VCM servo gantry system for motion tracking. Simulations and experimental results demonstrate good tracking performance, which outperform traditional control approaches. Copyright © 2016 ISA. Published by Elsevier Ltd. All rights reserved.

  11. Craniospinal treatment with IMRT multi-isocentric and image-guided linear accelerator based on Gantry

    International Nuclear Information System (INIS)

    Sanz Beltran, M.; Caballero Perea, B.; Rodriguez Rodriguez, C.; Arminio Diaz, E.; Lopez Fernandez, A.; Gomez Fervienza, J. R.; Crespo Diez, P.; Cantarero Valenzuela, N.; Alvarez Sanchez, M.; Martin Martin, G.; Gomez Fervienza, J. r.; Crespo Diez, P.; Cantarero Valenzuela, N.; Alvarez Sanchez, M.; Martin Martin, G.

    2011-01-01

    The objective is the realization of craniospinal treatment with a linear accelerator equipped with gantry based on MLC, carbon fiber table and Image Guided capability. The great length of treatment (patient l,80m in height) was a great difficulty for want of full length of the longitudinal movement of the table to adequately cover the PTV, plus free metallic screws fastening the head of the table extender preventing further incidents.

  12. Development of a MPPC-based prototype gantry for future MRI-PET scanners

    Science.gov (United States)

    Kurei, Y.; Kataoka, J.; Kato, T.; Fujita, T.; Ohshima, T.; Taya, T.; Yamamoto, S.

    2014-12-01

    We have developed a high spatial resolution, compact Positron Emission Tomography (PET) module designed for small animals and intended for use in magnetic resonance imaging (MRI) systems. This module consists of large-area, 4 × 4 ch MPPC arrays (S11830-3344MF; Hamamatsu Photonics K.K.) optically coupled with Ce-doped (Lu,Y)2(SiO4)O (Ce:LYSO) scintillators fabricated into 16 × 16 matrices of 0.5 × 0.5 mm2 pixels. We set the temperature sensor (LM73CIMK-0; National Semiconductor Corp.) at the rear of the MPPC acceptance surface, and apply optimum voltage to maintain the gain. The eight MPPC-based PET modules and coincidence circuits were assembled into a gantry arranged in a ring 90 mm in diameter to form the MPPC-based PET system. We have developed two types PET gantry: one made of non-magnetic metal and the other made of acrylonitrile butadiene styrene (ABS) resins. The PET gantry was positioned around the RF coil of the 4.7 T MRI system. We took an image of a point }22Na source under fast spin echo (FSE) and gradient echo (GE), in order to measure the interference between the MPPC-based PET and MRI. The spatial resolution of PET imaging in a transaxial plane of about 1 mm (FWHM) was achieved in all cases. Operating with PET made of ABS has no effect on MR images, while operating with PET made of non-magnetic metal has a significant detrimental effect on MR images. This paper describes our quantitative evaluations of PET images and MR images, and presents a more advanced version of the gantry for future MRI/DOI-PET systems.

  13. A comparison of robotic arm versus gantry linear accelerator stereotactic body radiation therapy for prostate cancer.

    Science.gov (United States)

    Avkshtol, Vladimir; Dong, Yanqun; Hayes, Shelly B; Hallman, Mark A; Price, Robert A; Sobczak, Mark L; Horwitz, Eric M; Zaorsky, Nicholas G

    2016-01-01

    Prostate cancer is the most prevalent cancer diagnosed in men in the United States besides skin cancer. Stereotactic body radiation therapy (SBRT; 6-15 Gy per fraction, up to 45 minutes per fraction, delivered in five fractions or less, over the course of approximately 2 weeks) is emerging as a popular treatment option for prostate cancer. The American Society for Radiation Oncology now recognizes SBRT for select low- and intermediate-risk prostate cancer patients. SBRT grew from the notion that high doses of radiation typical of brachytherapy could be delivered noninvasively using modern external-beam radiation therapy planning and delivery methods. SBRT is most commonly delivered using either a traditional gantry-mounted linear accelerator or a robotic arm-mounted linear accelerator. In this systematic review article, we compare and contrast the current clinical evidence supporting a gantry vs robotic arm SBRT for prostate cancer. The data for SBRT show encouraging and comparable results in terms of freedom from biochemical failure (>90% for low and intermediate risk at 5-7 years) and acute and late toxicity (6 MV). Finally, SBRT (particularly on a gantry) may also be more cost-effective than conventionally fractionated external-beam radiation therapy. Randomized controlled trials of SBRT using both technologies are underway.

  14. The effect of mandibular position and gantry angle on the evaluation of implant site with implant CT

    International Nuclear Information System (INIS)

    Lee, Sul Mi; An, Chang Hyeon; Heo, Min Suk; Lee, Sam Sun; Choi, Soon Chul; Park, Tae Won; Choi, Hang Moon

    2002-01-01

    The altered gantry angle during scanning for some multiplanar reconstruction CT program (CT/MPR) may cause distortion of the image. The aim of this study was to ascertain whether there is a image distortion in a reformatted image when the gantry and the object are equally inclined using ToothPix and DentaScan program.A resin block model with four cylindrical holes and a human dry mandible were used. Two MPR software packages, ToothPix and DentaScan program, were used for reformatted panoramic images. The block and the gantry were equally inclined at 0 degree, 15 degrees, and 30 degrees. With ToothPix program, a resin block model with empty holes and a dry mandible showed inclined images in the reformatted panoramic image. Increasing the gantry angle, the depth and inclination of the holes were increased in the reformatted central panoramic images. However, a resin block model with gutta percha in its holes and a dry mandible with a wire in its mandibular canal didn't show image distortion. With DentaScan program, image distortion was not seen in any situation. ToothPix program may distort the reformatted image when the gantry angle is not at zero degrees. However, with DentaScan program, the patient may be positioned comfortably and the gantry can be adjusted to the patient positioning.

  15. The Axial Curve Rotator.

    Science.gov (United States)

    Hunter, Walter M.

    This document contains detailed directions for constructing a device that mechanically produces the three-dimensional shape resulting from the rotation of any algebraic line or curve around either axis on the coordinate plant. The device was developed in response to student difficulty in visualizing, and thus grasping the mathematical principles…

  16. Effectiveness of MRI in rotator cuff injury

    International Nuclear Information System (INIS)

    Ohazama, Yuka

    1992-01-01

    To investigate the potential role of MR imaging in rotator cuf disorders, normal volunteers and patients with suspected rotator cuff injury were evaluated with a low field permanent magnet unit which had a wide gantry. MR findings of the patients were also compared with arthrography, subcromial bursography and operative findings. To establish optimal imaging technique and normal MR anatomy, 100 normal volunteers were examined. On proton density images, signal intensity of the rotator cuff tendon was low and homogenous, and that of rotator cuff muscles was intermediate. On T2 weighted images, signal intensity of muscles and tendon was decreased and that of joint effusion became brighter. In 38 patients with suspected rotator cuff injury, the signal intensity of the rotator cuff was increased to various degrees. In 21 of them, surgical correction was performed and 17 patients were followed with conservative treatment. MR imaging showed abnormalities in all 38 patients. Arthrography and bursography showed abnormalities in 28 out of 38 patients and 3 of 13 patients respectively. In 21 patients who underwent surgery, tear of the rotator cuff was confirmed, and discrepancies in MR and operative findings existed in 8 patients. In 2 patients, no tear was found in the other examinations, and it was suspected to be horizontal tear or degeneration in the substance of the muscle. MR imaging contributes to diagnosis and treatment planning in patients with suspected rotator cuff injury. (author)

  17. Overview of CMS robotic silicon module assembly hardware based on Aerotech Gantry Positioning system.

    CERN Multimedia

    Honma, Alan

    1999-01-01

    The goal of the robotic silicon module assembly pilot project is to fully automate the gluing and pick and placement of silicon sensors and front-end hybrid onto a carbon-fibre frame. The basis for thesystem is the Aerotech Gantry Positioning System (AGS10000) machineshown in the centre of the picture. To the left is the PC which contains the controller card and runs the user interface. To the rightis the rack of associated electronics which interfaces with the CERNbuilt tooling and vacuum chuck system.

  18. Rotation sensor switch

    International Nuclear Information System (INIS)

    Sevec, J.B.

    1978-01-01

    A protective device to provide a warning if a piece of rotating machinery slows or stops is comprised of a pair of hinged weights disposed to rotate on a rotating shaft of the equipment. When the equipment is rotating, the weights remain in a plane essentially perpendicular to the shaft and constitute part of an electrical circuit that is open. When the shaft slows or stops, the weights are attracted to a pair of concentric electrically conducting disks disposed in a plane perpendicular to the shaft and parallel to the plane of the weights when rotating. A disk magnet attracts the weights to the electrically conducting plates and maintains the electrical contact at the plates to complete an electrical circuit that can then provide an alarm signal

  19. Hadron cancer therapy complex using nonscaling fixed field alternating gradient accelerator and gantry design

    Directory of Open Access Journals (Sweden)

    E. Keil

    2007-05-01

    Full Text Available Nonscaling fixed field alternating gradient (FFAG rings for cancer hadron therapy offer reduced physical aperture and large dynamic aperture as compared to scaling FFAGs. The variation of tune with energy implies the crossing of resonances during acceleration. Our design avoids intrinsic resonances, although imperfection resonances must be crossed. We consider a system of three nonscaling FFAG rings for cancer therapy with 250 MeV protons and 400   MeV/u carbon ions. Hadrons are accelerated in a common radio frequency quadrupole and linear accelerator, and injected into the FFAG rings at v/c=0.1294. H^{+}/C^{6+} ions are accelerated in the two smaller/larger rings to 31 and 250  MeV/68.8 and 400   MeV/u kinetic energy, respectively. The lattices consist of doublet cells with a straight section for rf cavities. The gantry with triplet cells accepts the whole required momentum range at fixed field. This unique design uses either high-temperature superconductors or superconducting magnets reducing gantry magnet size and weight. Elements with a variable field at the beginning and at the end set the extracted beam at the correct position for a range of energies.

  20. Robotically Assembled Aerospace Structures: Digital Material Assembly using a Gantry-Type Assembler

    Science.gov (United States)

    Trinh, Greenfield; Copplestone, Grace; O'Connor, Molly; Hu, Steven; Nowak, Sebastian; Cheung, Kenneth; Jenett, Benjamin; Cellucci, Daniel

    2017-01-01

    This paper evaluates the development of automated assembly techniques for discrete lattice structures using a multi-axis gantry type CNC machine. These lattices are made of discrete components called "digital materials." We present the development of a specialized end effector that works in conjunction with the CNC machine to assemble these lattices. With this configuration we are able to place voxels at a rate of 1.5 per minute. The scalability of digital material structures due to the incremental modular assembly is one of its key traits and an important metric of interest. We investigate the build times of a 5x5 beam structure on the scale of 1 meter (325 parts), 10 meters (3,250 parts), and 30 meters (9,750 parts). Utilizing the current configuration with a single end effector, performing serial assembly with a globally fixed feed station at the edge of the build volume, the build time increases according to a scaling law of n4, where n is the build scale. Build times can be reduced significantly by integrating feed systems into the gantry itself, resulting in a scaling law of n3. A completely serial assembly process will encounter time limitations as build scale increases. Automated assembly for digital materials can assemble high performance structures from discrete parts, and techniques such as built in feed systems, parallelization, and optimization of the fastening process will yield much higher throughput.

  1. Rotating flow

    CERN Document Server

    Childs, Peter R N

    2010-01-01

    Rotating flow is critically important across a wide range of scientific, engineering and product applications, providing design and modeling capability for diverse products such as jet engines, pumps and vacuum cleaners, as well as geophysical flows. Developed over the course of 20 years' research into rotating fluids and associated heat transfer at the University of Sussex Thermo-Fluid Mechanics Research Centre (TFMRC), Rotating Flow is an indispensable reference and resource for all those working within the gas turbine and rotating machinery industries. Traditional fluid and flow dynamics titles offer the essential background but generally include very sparse coverage of rotating flows-which is where this book comes in. Beginning with an accessible introduction to rotating flow, recognized expert Peter Childs takes you through fundamental equations, vorticity and vortices, rotating disc flow, flow around rotating cylinders and flow in rotating cavities, with an introduction to atmospheric and oceanic circul...

  2. Rotational seismology

    Science.gov (United States)

    Lee, William H K.

    2016-01-01

    Rotational seismology is an emerging study of all aspects of rotational motions induced by earthquakes, explosions, and ambient vibrations. It is of interest to several disciplines, including seismology, earthquake engineering, geodesy, and earth-based detection of Einstein’s gravitation waves.Rotational effects of seismic waves, together with rotations caused by soil–structure interaction, have been observed for centuries (e.g., rotated chimneys, monuments, and tombstones). Figure 1a shows the rotated monument to George Inglis observed after the 1897 Great Shillong earthquake. This monument had the form of an obelisk rising over 19 metres high from a 4 metre base. During the earthquake, the top part broke off and the remnant of some 6 metres rotated about 15° relative to the base. The study of rotational seismology began only recently when sensitive rotational sensors became available due to advances in aeronautical and astronomical instrumentations.

  3. 21 CFR 890.5180 - Manual patient rotation bed.

    Science.gov (United States)

    2010-04-01

    ...) MEDICAL DEVICES PHYSICAL MEDICINE DEVICES Physical Medicine Therapeutic Devices § 890.5180 Manual patient rotation bed. (a) Identification. A manual patient rotation bed is a device that turns a patient who is... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Manual patient rotation bed. 890.5180 Section 890...

  4. 21 CFR 890.5225 - Powered patient rotation bed.

    Science.gov (United States)

    2010-04-01

    ...) MEDICAL DEVICES PHYSICAL MEDICINE DEVICES Physical Medicine Therapeutic Devices § 890.5225 Powered patient rotation bed. (a) Identification. A powered patient rotation bed is a device that turns a patient who is... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Powered patient rotation bed. 890.5225 Section 890...

  5. Guide device

    International Nuclear Information System (INIS)

    Brammer, C.M. Jr.

    1977-01-01

    Disclosed is a fuel handling guide tube centering device for use in nuclear reactors during fuel assembly handling operations. The device comprises an outer ring secured to the flange of a nuclear reactor pressure vessel, a rotatable table rotatably coupled to the outer ring, and a plurality of openings through the table. Truncated locating cones are positioned in each of the openings in the table, and the locating cones center the guide tube during fuel handling operations. The openings in the table are located such that each fuel assembly in the nuclear core may be aligned with one of the openings by a suitable rotation of the table. The locating cones thereby provide alignment between the fuel handling mechanism located in the guide tube and the individual fuel assemblies of the cone. The need for a device to provide alignment is especially critical for floating nuclear power plants, where wave motion may exist during fuel handling operations. 5 claims, 4 figures

  6. Precedent Research on Compact Laser-plasma based Gantry for Cancer Therapy

    International Nuclear Information System (INIS)

    Hee, Park Seong; Jeong, Young Uk; Lee, Ki Tae; Kim, Kyung Nam; Cha, Young Ho

    2012-03-01

    This is the precedent R and D to develop the technology of next generation compact particle cancer treatment system based on laser-plasma interaction and to deduce a big project. The subject of this project are the survey of application technology of laser-plasma based particle beam and the design of compact laser-plasma based gantry. The survey of characteristic of particle beam for cancer therapy and present status can be adapted to develop new system. The comparison between particle beams from the existing system and new one based on laser-plasma acceleration will be important to new design and design optimization. The project includes design of multi-dimensional laser transfer beamline, minimization of laser-plasma acceleration chamber, design of effective energy separation/selection system, and radiation safety and local shielding

  7. Dosimetric Comparison of Manual and Beam Angle Optimization of Gantry Angles in IMRT

    International Nuclear Information System (INIS)

    Srivastava, Shiv P.; Das, Indra J.; Kumar, Arvind; Johnstone, Peter A.S.

    2011-01-01

    Dosimetric comparison of manual beam angle selection (MBS) and beam angle optimization (BAO) for IMRT plans is investigated retrospectively for 15 head and neck and prostate patients. The head and neck and prostate had planning target volumes (PTVs) ranging between 96.0 and 319.9 cm 3 and 153.6 and 321.3 cm 3 , whereas OAR ranged between 8.3 and 47.8 cm 3 and 68.3 and 469.2 cm 3 , respectively. In MBS, a standard coplanar 7-9 fields equally spaced gantry angles were used. In BAO, the selection of gantry angle was optimized by the algorithm for the same number of beams. The optimization and dose-volume constraints were kept the same for both techniques. Treatment planning was performed on the Eclipse treatment planning system. Our results showed that the dose-volume histogram for PTV are nearly identical in both techniques but BAO provided superior sparing of the organs at risk compared with the MBS. Also, MBS produced statistically significant higher monitor units (MU) and segments than the BAO; 13.1 ± 6.6% (p = 0.012) and 10.4 ± 13.6% (p = 0.140), and 14.6 ± 5.6% (p = 1.003E-5) and 12.6 ± 7.4% (p = 0.76E-3) for head and neck and prostate cases, respectively. The reduction in MU translates into the reduction in total body and integral dose. It is concluded that BAO provides advantage over MBS for most intenisty-modulated radiation therapy cases.

  8. Global rotation

    International Nuclear Information System (INIS)

    Rosquist, K.

    1980-01-01

    Global rotation in cosmological models is defined on an observational basis. A theorem is proved saying that, for rigid motion, the global rotation is equal to the ordinary local vorticity. The global rotation is calculated in the space-time homogeneous class III models, with Godel's model as a special case. It is shown that, with the exception of Godel's model, the rotation in these models becomes infinite for finite affine parameter values. In some directions the rotation changes sign and becomes infinite in a direction opposite to the local vorticity. The points of infinite rotation are identified as conjugate points along the null geodesics. The physical interpretation of the infinite rotation is discussed, and a comparison with the behaviour of the area distance at conjugate points is given. (author)

  9. Quality assurance system to correct for errors arising from couch rotation in linac-based stereotactic radiosurgery

    International Nuclear Information System (INIS)

    Brezovich, Ivan A.; Pareek, Prem N.; Plott, W. Eugene; Jennelle, Richard L. S.

    1997-01-01

    Purpose: The purpose of this project was the development of a quality assurance (QA) system that would provide geographically accurate targeting for linac-based stereotactic radiosurgery (LBSR). Methods and Materials: The key component of our QA system is a novel device (Alignment Tool) for expedient measurement of gantry and treatment table excursions (wobble) during rotation. The Alignment Tool replaces the familiar pencil-shaped pointers with a ball pointer that is used with the field light of the accelerator to indicate alignment of beam and target. Wobble is measured prior to each patient treatment and analyzed together with the BRW coordinates of the target by a spreadsheet. The corrections required to compensate for any imprecision are identified, and a printout generated indicating the floor stand coordinates for each couch angle used to place the target at isocenter. Results: The Alignment Tool has an inherent accuracy of measurement better than 0.1 mm. The overall targeting error of our QA method, found by evaluating 177 target simulator films of 55 foci in 40 randomly selected patients, was 0.47 ± 0.23 mm. The Alignment Tool was also valuable during installation of the floor stand and a supplemental collimator for the accelerator. Conclusions: The QA procedure described allows accurate targeting in LBSR, even when couch rotation is imprecise. The Alignment Tool can facilitate the installation of any stereotactic irradiation system, and can be useful for annual QA checks as well as in the installation and commissioning of new accelerators

  10. VMAT optimization with dynamic collimator rotation.

    Science.gov (United States)

    Lyu, Qihui; O'Connor, Daniel; Ruan, Dan; Yu, Victoria; Nguyen, Dan; Sheng, Ke

    2018-04-16

    Although collimator rotation is an optimization variable that can be exploited for dosimetric advantages, existing Volumetric Modulated Arc Therapy (VMAT) optimization uses a fixed collimator angle in each arc and only rotates the collimator between arcs. In this study, we develop a novel integrated optimization method for VMAT, accounting for dynamic collimator angles during the arc motion. Direct Aperture Optimization (DAO) for Dynamic Collimator in VMAT (DC-VMAT) was achieved by adding to the existing dose fidelity objective an anisotropic total variation term for regulating the fluence smoothness, a binary variable for forming simple apertures, and a group sparsity term for controlling collimator rotation. The optimal collimator angle for each beam angle was selected using the Dijkstra's algorithm, where the node costs depend on the estimated fluence map at the current iteration and the edge costs account for the mechanical constraints of multi-leaf collimator (MLC). An alternating optimization strategy was implemented to solve the DAO and collimator angle selection (CAS). Feasibility of DC-VMAT using one full-arc with dynamic collimator rotation was tested on a phantom with two small spherical targets, a brain, a lung and a prostate cancer patient. The plan was compared against a static collimator VMAT (SC-VMAT) plan using three full arcs with 60 degrees of collimator angle separation in patient studies. With the same target coverage, DC-VMAT achieved 20.3% reduction of R50 in the phantom study, and reduced the average max and mean OAR dose by 4.49% and 2.53% of the prescription dose in patient studies, as compared with SC-VMAT. The collimator rotation co-ordinated with the gantry rotation in DC-VMAT plans for deliverability. There were 13 beam angles in the single-arc DC-VMAT plan in patient studies that requires slower gantry rotation to accommodate multiple collimator angles. The novel DC-VMAT approach utilizes the dynamic collimator rotation during arc

  11. Flexural pivot device

    International Nuclear Information System (INIS)

    Flaherty, Robert.

    1986-01-01

    A flexural pivot device or rotational actuator comprises first and sceond tubular members connected by flexural members of shape-memory-alloy. These are curved in the austenitic phase at a first temperature and after cooling to the martensitic phase are flattened. On heating one of the flexural members, it bends causing relative rotation of the tubular members. Heating of another member can produce opposite rotation. Heating is electrical or by hot gas. The device may be used in a nuclear reactor. (author)

  12. Rotating Wavepackets

    Science.gov (United States)

    Lekner, John

    2008-01-01

    Any free-particle wavepacket solution of Schrodinger's equation can be converted by differentiations to wavepackets rotating about the original direction of motion. The angular momentum component along the motion associated with this rotation is an integral multiple of [h-bar]. It is an "intrinsic" angular momentum: independent of origin and…

  13. Rotating dryer

    International Nuclear Information System (INIS)

    Noe, C.

    1984-01-01

    Products to dry are introduced inside a rotating tube placed in an oven, the cross section of the tube is an arc of spiral. During clockwise rotation of the tube products are maintained inside and mixed, during anticlockwise products are removed. Application is made to drying of radioactive wastes [fr

  14. Rotation, Stability and Transport

    Energy Technology Data Exchange (ETDEWEB)

    Connor, J. W.

    2007-07-01

    scales. Some current theoretical challenges, such as reconciling sheared plasma rotation and ballooning mode theory, will be mentioned. Predicted effects on plasma stability, both for macroscopic MHD modes and for drift waves, as well as for transport will be discussed. The theoretical modelling of rotation and its effects will be linked to experimental observations on various magnetic confinement devices. (Author)

  15. Craniospinal treatment with IMRT multi-isocentric and image-guided linear accelerator based on Gantry; Tratamiento craneoespinal multi-isocentrico con IMRT y guiado por imagen en acelerador lineal basado en Gantry

    Energy Technology Data Exchange (ETDEWEB)

    Sanz Beltran, M.; Caballero Perea, B.; Rodriguez Rodriguez, C.; Arminio Diaz, E.; Lopez Fernandez, A.; Gomez Fervienza, J. R.; Crespo Diez, P.; Cantarero Valenzuela, N.; Alvarez Sanchez, M.; Martin Martin, G.; Gomez Fervienza, J. r.; Crespo Diez, P.; Cantarero Valenzuela, N.; Alvarez Sanchez, M.; Martin Martin, G.

    2011-07-01

    The objective is the realization of craniospinal treatment with a linear accelerator equipped with gantry based on MLC, carbon fiber table and Image Guided capability. The great length of treatment (patient l, 80m in height) was a great difficulty for want of full length of the longitudinal movement of the table to adequately cover the PTV, plus free metallic screws fastening the head of the table extender preventing further incidents.

  16. A rotating arc plasma invertor

    International Nuclear Information System (INIS)

    Reusch, M.F.; Jayaram, K.

    1987-02-01

    A device is described for the inversion of direct current to alternating current. The main feature is the use of a rotating plasma arc in crossed electric and magnetic fields as a switch. This device may provide an economic alternative to other inversion methods in some circumstances

  17. Improved scintimammography using a high-resolution camera mounted on an upright mammography gantry

    International Nuclear Information System (INIS)

    Itti, Emmanuel; Patt, Bradley E.; Diggles, Linda E.; MacDonald, Lawrence; Iwanczyk, Jan S.; Mishkin, Fred S.; Khalkhali, Iraj

    2003-01-01

    99m Tc-sestamibi scintimammography (SMM) is a useful adjunct to conventional X-ray mammography (XMM) for the assessment of breast cancer. An increasing number of studies has emphasized fair sensitivity values for the detection of tumors >1 cm, compared to XMM, particularly in situations where high glandular breast densities make mammographic interpretation difficult. In addition, SMM has demonstrated high specificity for cancer, compared to various functional and anatomic imaging modalities. However, large field-of-view (FOV) gamma cameras are difficult to position close to the breasts, which decreases spatial resolution and subsequently, the sensitivity of detection for tumors 2 FOV and an array of 2x2x6 mm 3 discrete crystals coupled to a photon-sensitive photomultiplier tube readout. This camera is mounted on a mammography gantry allowing upright imaging, medial positioning and use of breast compression. Preliminary data indicates significant enhancement of spatial resolution by comparison with standard imaging in the first 10 patients. Larger series will be needed to conclude on sensitivity/specificity issues

  18. Improved scintimammography using a high-resolution camera mounted on an upright mammography gantry

    Energy Technology Data Exchange (ETDEWEB)

    Itti, Emmanuel; Patt, Bradley E.; Diggles, Linda E.; MacDonald, Lawrence; Iwanczyk, Jan S.; Mishkin, Fred S.; Khalkhali, Iraj E-mail: nephrad@aol.com

    2003-01-21

    {sup 99m}Tc-sestamibi scintimammography (SMM) is a useful adjunct to conventional X-ray mammography (XMM) for the assessment of breast cancer. An increasing number of studies has emphasized fair sensitivity values for the detection of tumors >1 cm, compared to XMM, particularly in situations where high glandular breast densities make mammographic interpretation difficult. In addition, SMM has demonstrated high specificity for cancer, compared to various functional and anatomic imaging modalities. However, large field-of-view (FOV) gamma cameras are difficult to position close to the breasts, which decreases spatial resolution and subsequently, the sensitivity of detection for tumors <1 cm. New dedicated detectors featuring small FOV and increased spatial resolution have recently been developed. In this setting, improvement in tumor detection sensitivity, particularly with regard to small cancers is expected. At Division of Nuclear Medicine, Harbor-UCLA Medical Center, we have performed over 2000 SMM within the last 9 years. We have recently used a dedicated breast camera (LumaGEM) featuring a 12.8x12.8 cm{sup 2} FOV and an array of 2x2x6 mm{sup 3} discrete crystals coupled to a photon-sensitive photomultiplier tube readout. This camera is mounted on a mammography gantry allowing upright imaging, medial positioning and use of breast compression. Preliminary data indicates significant enhancement of spatial resolution by comparison with standard imaging in the first 10 patients. Larger series will be needed to conclude on sensitivity/specificity issues.

  19. Technical Note: Kinect V2 surface filtering during gantry motion for radiotherapy applications.

    Science.gov (United States)

    Nazir, Souha; Rihana, Sandy; Visvikis, Dimitris; Fayad, Hadi

    2018-04-01

    In radiotherapy, the Kinect V2 camera, has recently received a lot of attention concerning many clinical applications including patient positioning, respiratory motion tracking, and collision detection during the radiotherapy delivery phase. However, issues associated with such applications are related to some materials and surfaces reflections generating an offset in depth measurements especially during gantry motion. This phenomenon appears in particular when the collimator surface is observed by the camera; resulting in erroneous depth measurements, not only in Kinect surfaces itself, but also as a large peak when extracting a 1D respiratory signal from these data. In this paper, we proposed filtering techniques to reduce the noise effect in the Kinect-based 1D respiratory signal, using a trend removal filter, and in associated 2D surfaces, using a temporal median filter. Filtering process was validated using a phantom, in order to simulate a patient undergoing radiotherapy treatment while having the ground truth. Our results indicate a better correlation between the reference respiratory signal and its corresponding filtered signal (Correlation coefficient of 0.76) than that of the nonfiltered signal (Correlation coefficient of 0.13). Furthermore, surface filtering results show a decrease in the mean square distance error (85%) between the reference and the measured point clouds. This work shows a significant noise compensation and surface restitution after surface filtering and therefore a potential use of the Kinect V2 camera for different radiotherapy-based applications, such as respiratory tracking and collision detection. © 2018 American Association of Physicists in Medicine.

  20. Design and testing of a rotating, cooled device for extra-corporate treatment of liver cancer by BNCT in the epithermal neutron beam at the HFR Petten

    International Nuclear Information System (INIS)

    Moss, Ray; Nievaart, Sander; Pott, Lucien; Wittig, Andrea; Sauerwein, Wolfgang

    2006-01-01

    As part of the joint project on extra-corporal treatment of liver cancer by BNCT between JRC Petten and the University Hospital Essen, a facility has been designed and built to contain the liver during its irradiation treatment at the HFR Petten. The design consists of a rotating spheroid shaped PMMA holder, manufactured to open at the equator and closed by screwing together, surrounded by PMMA and graphite blocks. A validation exercise has been performed regarding both the nuclear conditions and the physical conditions. For the former, activation foil sets of Au, Cu and Mn, were irradiated at positions inside the liver holder filled with water, whilst a second measurement campaign has been performed using gel dosimetry. For the physical test, it is required to operate (rotate) the facility for up to 4 hours and to maintain the liver at approximately 4degC. The latter test was performed using 'cold gun sprays' that inject cold air near the liver holder. Both the nuclear and physical validation tests were performed successfully. (author)

  1. Design of 6 MeV X-band electron linac for dual-head gantry radiotherapy system

    Science.gov (United States)

    Shin, Seung-wook; Lee, Seung-Hyun; Lee, Jong-Chul; Kim, Huisu; Ha, Donghyup; Ghergherehchi, Mitra; Chai, Jongseo; Lee, Byung-no; Chae, Moonsik

    2017-12-01

    A compact 6 MeV electron linac is being developed at Sungkyunkwan University, in collaboration with the Korea atomic energy research institute (KAERI). The linac will be used as an X-ray source for a dual-head gantry radiotherapy system. X-band technology has been employed to satisfy the size requirement of the dual-head gantry radiotherapy machine. Among the several options available, we selected a pi/2-mode, standing-wave, side-coupled cavity. This choice of radiofrequency (RF) cavity design is intended to enhance the shunt impedance of each cavity in the linac. An optimum structure of the RF cavity with a high-performance design was determined by applying a genetic algorithm during the optimization procedure. This paper describes the detailed design process for a single normal RF cavity and the entire structure, including the RF power coupler and coupling cavity, as well as the beam dynamics results.

  2. Optimal Performance of a Nonlinear Gantry Crane System via Priority-based Fitness Scheme in Binary PSO Algorithm

    International Nuclear Information System (INIS)

    Jaafar, Hazriq Izzuan; Ali, Nursabillilah Mohd; Selamat, Nur Asmiza; Kassim, Anuar Mohamed; Mohamed, Z; Abidin, Amar Faiz Zainal; Jamian, J J

    2013-01-01

    This paper presents development of an optimal PID and PD controllers for controlling the nonlinear gantry crane system. The proposed Binary Particle Swarm Optimization (BPSO) algorithm that uses Priority-based Fitness Scheme is adopted in obtaining five optimal controller gains. The optimal gains are tested on a control structure that combines PID and PD controllers to examine system responses including trolley displacement and payload oscillation. The dynamic model of gantry crane system is derived using Lagrange equation. Simulation is conducted within Matlab environment to verify the performance of system in terms of settling time (Ts), steady state error (SSE) and overshoot (OS). This proposed technique demonstrates that implementation of Priority-based Fitness Scheme in BPSO is effective and able to move the trolley as fast as possible to the various desired position

  3. Reducing 4DCBCT imaging time and dose: the first implementation of variable gantry speed 4DCBCT on a linear accelerator.

    Science.gov (United States)

    O'Brien, Ricky T; Stankovic, Uros; Sonke, Jan-Jakob; Keall, Paul J

    2017-06-07

    Four dimensional cone beam computed tomography (4DCBCT) uses a constant gantry speed and imaging frequency that are independent of the patient's breathing rate. Using a technique called respiratory motion guided 4DCBCT (RMG-4DCBCT), we have previously demonstrated that by varying the gantry speed and imaging frequency, in response to changes in the patient's real-time respiratory signal, the imaging dose can be reduced by 50-70%. RMG-4DCBCT optimally computes a patient specific gantry trajectory to eliminate streaking artefacts and projection clustering that is inherent in 4DCBCT imaging. The gantry trajectory is continuously updated as projection data is acquired and the patient's breathing changes. The aim of this study was to realise RMG-4DCBCT for the first time on a linear accelerator. To change the gantry speed in real-time a potentiometer under microcontroller control was used to adjust the current supplied to an Elekta Synergy's gantry motor. A real-time feedback loop was developed on the microcontroller to modulate the gantry speed and projection acquisition in response to the real-time respiratory signal so that either 40, RMG-4DCBCT 40 , or 60, RMG-4DCBCT 60 , uniformly spaced projections were acquired in 10 phase bins. Images of the CIRS dynamic Thorax phantom were acquired with sinusoidal breathing periods ranging from 2 s to 8 s together with two breathing traces from lung cancer patients. Image quality was assessed using the contrast to noise ratio (CNR) and edge response width (ERW). For the average patient, with a 3.8 s breathing period, the imaging time and image dose were reduced by 37% and 70% respectively. Across all respiratory rates, RMG-4DCBCT 40 had a CNR in the range of 6.5 to 7.5, and RMG-4DCBCT 60 had a CNR between 8.7 and 9.7, indicating that RMG-4DCBCT allows consistent and controllable CNR. In comparison, the CNR for conventional 4DCBCT drops from 20.4 to 6.2 as the breathing rate increases from 2 s to 8 s. With RMG-4DCBCT

  4. PLASMA DEVICE

    Science.gov (United States)

    Gow, J.D.; Wilcox, J.M.

    1961-12-26

    A device is designed for producing and confining highenergy plasma from which neutrons are generated in copious quantities. A rotating sheath of electrons is established in a radial electric field and axial magnetic field produced within the device. The electron sheath serves as a strong ionizing medium to gas introdueed thereto and also functions as an extremely effective heating mechanism to the resulting plasma. In addition, improved confinement of the plasma is obtained by ring magnetic mirror fields produced at the ends of the device. Such ring mirror fields are defined by the magnetic field lines at the ends of the device diverging radially outward from the axis of the device and thereafter converging at spatial annular surfaces disposed concentrically thereabout. (AFC)

  5. More than 10 years experience of beam monitoring with the Gantry 1 spot scanning proton therapy facility at PSI

    International Nuclear Information System (INIS)

    Lin Shixiong; Boehringer, Terence; Coray, Adolf; Grossmann, Martin; Pedroni, Eros

    2009-01-01

    Purpose: The beam monitoring equipments developed for the first PSI spot scanning proton therapy facility, Gantry 1, have been successfully used for more than 10 years. The purpose of this article is to summarize the author's experience in the beam monitoring technique for dynamic proton scanning. Methods: The spot dose delivery and verification use two independent beam monitoring and computer systems. In this article, the detector construction, electronic system, dosimetry, and quality assurance results are described in detail. The beam flux monitor is calibrated with a Faraday cup. The beam position monitoring is realized by measuring the magnetic fields of deflection magnets with Hall probes before applying the spot and by checking the beam position and width with an ionization strip chamber after the spot delivery. Results: The results of thimble ionization chamber dosimetry measurements are reproducible (with a mean deviation of less than 1% and a standard deviation of 1%). The resolution in the beam position measurement is of the order of a tenth of a millimeter. The tolerance of the beam position delivery and monitoring during scanning is less than 1.5 mm. Conclusions: The experiences gained with the successful operation of Gantry 1 represent a unique and solid background for the development of a new system, Gantry 2, in order to perform new advanced scanning techniques.

  6. More than 10 years experience of beam monitoring with the Gantry 1 spot scanning proton therapy facility at PSI

    Energy Technology Data Exchange (ETDEWEB)

    Lin Shixiong; Boehringer, Terence; Coray, Adolf; Grossmann, Martin; Pedroni, Eros [Center for Proton Therapy, Paul Scherrer Institute, CH-5232 Villigen PSI (Switzerland)

    2009-11-15

    Purpose: The beam monitoring equipments developed for the first PSI spot scanning proton therapy facility, Gantry 1, have been successfully used for more than 10 years. The purpose of this article is to summarize the author's experience in the beam monitoring technique for dynamic proton scanning. Methods: The spot dose delivery and verification use two independent beam monitoring and computer systems. In this article, the detector construction, electronic system, dosimetry, and quality assurance results are described in detail. The beam flux monitor is calibrated with a Faraday cup. The beam position monitoring is realized by measuring the magnetic fields of deflection magnets with Hall probes before applying the spot and by checking the beam position and width with an ionization strip chamber after the spot delivery. Results: The results of thimble ionization chamber dosimetry measurements are reproducible (with a mean deviation of less than 1% and a standard deviation of 1%). The resolution in the beam position measurement is of the order of a tenth of a millimeter. The tolerance of the beam position delivery and monitoring during scanning is less than 1.5 mm. Conclusions: The experiences gained with the successful operation of Gantry 1 represent a unique and solid background for the development of a new system, Gantry 2, in order to perform new advanced scanning techniques.

  7. A molecularly imprinted polymer as the sorptive phase immobilized in a rotating disk extraction device for the determination of diclofenac and mefenamic acid in wastewater

    International Nuclear Information System (INIS)

    Manzo, Valentina; Ulisse, Karla; Rodríguez, Inés; Pereira, Eduardo; Richter, Pablo

    2015-01-01

    The microextraction of diclofenac and mefenamic acid from water samples was performed by using rotating disk sorptive extraction (RDSE) with molecularly imprinted polymer (MIP) as the sorptive phase. The MIP was synthesized from the monomer 1-vinylimidazol (VI) together with the cross-linker divinylbenzene (DVB) using diphenylamine as the template molecule. Scanning electron microscopy (SEM) analyses of the MIP revealed clusters of spherical particles having a narrow size distribution, with diameters of approximately 1 μm. The optimized extraction conditions involved a disk rotation velocity of 3000 rpm, an extraction time of 120 min, a sample volume of 50 mL, and a sample pH of 2 as well as 25 mg of MIP immobilized in the disk. Desorption of the extracted analytes was performed with 5 mL of methanol for 10 min. Analysis by gas chromatography-mass spectrometry (GC–MS) was carried out after derivatization of the analytes with N-tert-butyldimethylsilyl-N-methyltrifluoroacetamide (MTBSTFA). Nonmolecularly imprinted polymer (NIP) was also synthesized for comparison. It was observed that under the same conditions, MIP extracted significantly more NSAIDs containing diphenylamine (or part of this molecule) in their structure than NIP. Higher significant differences between MIP and NIP were observed for diclofenac, mefenamic acid and paracetamol, clearly indicating the effect of the template on the extraction. Recoveries of the method were between 100 and 112%, with relative standard deviations of 5–6%. The limits of detection were between 60 and 223 ng L −1 . Water samples from a wastewater treatment plant (WWTP) of Santiago de Chile, were found to contain concentrations of these acidic drugs between 1.6 and 4.3 μg L −1 and between 1.4 and 3.3 μg L −1 in the influent and effluent, respectively. - Highlights: • A MIP immobilized in a rotating disk sucessfully extracts NSAIDs from wastewater. • MIP had remarkably superior binding properties

  8. A molecularly imprinted polymer as the sorptive phase immobilized in a rotating disk extraction device for the determination of diclofenac and mefenamic acid in wastewater

    Energy Technology Data Exchange (ETDEWEB)

    Manzo, Valentina; Ulisse, Karla [Department of Inorganic and Analytical Chemistry, Faculty of Chemical and Pharmaceutical Sciences, University of Chile, P.O. Box 653, Santiago (Chile); Rodríguez, Inés [Department of Analytical and Inorganic Chemistry, Faculty of Chemical Sciences, University of Concepción (Chile); Pereira, Eduardo, E-mail: epereira@udec.cl [Department of Analytical and Inorganic Chemistry, Faculty of Chemical Sciences, University of Concepción (Chile); Richter, Pablo, E-mail: prichter@ciq.uchile.cl [Department of Inorganic and Analytical Chemistry, Faculty of Chemical and Pharmaceutical Sciences, University of Chile, P.O. Box 653, Santiago (Chile)

    2015-08-19

    The microextraction of diclofenac and mefenamic acid from water samples was performed by using rotating disk sorptive extraction (RDSE) with molecularly imprinted polymer (MIP) as the sorptive phase. The MIP was synthesized from the monomer 1-vinylimidazol (VI) together with the cross-linker divinylbenzene (DVB) using diphenylamine as the template molecule. Scanning electron microscopy (SEM) analyses of the MIP revealed clusters of spherical particles having a narrow size distribution, with diameters of approximately 1 μm. The optimized extraction conditions involved a disk rotation velocity of 3000 rpm, an extraction time of 120 min, a sample volume of 50 mL, and a sample pH of 2 as well as 25 mg of MIP immobilized in the disk. Desorption of the extracted analytes was performed with 5 mL of methanol for 10 min. Analysis by gas chromatography-mass spectrometry (GC–MS) was carried out after derivatization of the analytes with N-tert-butyldimethylsilyl-N-methyltrifluoroacetamide (MTBSTFA). Nonmolecularly imprinted polymer (NIP) was also synthesized for comparison. It was observed that under the same conditions, MIP extracted significantly more NSAIDs containing diphenylamine (or part of this molecule) in their structure than NIP. Higher significant differences between MIP and NIP were observed for diclofenac, mefenamic acid and paracetamol, clearly indicating the effect of the template on the extraction. Recoveries of the method were between 100 and 112%, with relative standard deviations of 5–6%. The limits of detection were between 60 and 223 ng L{sup −1}. Water samples from a wastewater treatment plant (WWTP) of Santiago de Chile, were found to contain concentrations of these acidic drugs between 1.6 and 4.3 μg L{sup −1} and between 1.4 and 3.3 μg L{sup −1} in the influent and effluent, respectively. - Highlights: • A MIP immobilized in a rotating disk sucessfully extracts NSAIDs from wastewater. • MIP had remarkably superior binding

  9. Rotatable seal assembly

    International Nuclear Information System (INIS)

    Garibaldi, J.L.; Logan, C.M.

    1982-01-01

    An assembly is provided for rotatably supporting a rotor on a stator so that vacuum chambers in the rotor and stator remain in communication while the chambers are sealed from ambient air, which enables the use of a ball bearing or the like to support most of the weight of the rotor. The apparatus includes a seal device mounted on the rotor to rotate therewith, but shiftable in position on the rotor while being sealed to the rotor as by an oring. The seal device has a flat face that is biased towards a flat face on the stator, and pressurized air is pumped between the faces to prevent contact between them while spacing them a small distance apart to avoid the inflow of large amounts of air between the faces and into the vacuum chambers

  10. Optical illusions induced by rotating medium

    Science.gov (United States)

    Zang, XiaoFei; Huang, PengCheng; Zhu, YiMing

    2018-03-01

    Different from the traditional single-function electromagnetic wave rotators (rotate the electromagnetic wavefronts), we propose that rotating medium can be extended to optical illusions such as breaking the diffraction limit and overlapping illusion. Furthermore, the homogeneous but anisotropic rotating medium is simplified by homogeneous and isotropic positive-index materials according to the effective medium theory, which is helpful for future device fabrication. Finite element simulations for the two-dimensional case are performed to demonstrate these properties.

  11. Wave-Driven Rotation In Centrifugal Mirrors

    International Nuclear Information System (INIS)

    Fetterman, Abraham J.; Fisch, Nathaniel J.

    2011-01-01

    Centrifugal mirrors use supersonic rotation to provide axial confinement and enhanced stability. Usually the rotation is produced using electrodes, but these electrodes have limited the rotation to the Alfven critical ionization velocity, which is too slow to be useful for fusion. Instead, the rotation could be produced using radio frequency waves. A fixed azimuthal ripple is a simple and efficient wave that could produce rotation by harnessing alpha particle energy. This is an extension of the alpha channeling effect. The alpha particle power and efficiency in a simulated devices is sufficient to produce rotation without external energy input. By eliminating the need for electrodes, this opens new opportunities for centrifugal traps.

  12. Predictive Parameters of Symptomatic Hematochezia Following 5-Fraction Gantry-Based SABR in Prostate Cancer

    International Nuclear Information System (INIS)

    Musunuru, Hima Bindu; Davidson, Melanie; Cheung, Patrick; Vesprini, Danny; Liu, Stanley; Chung, Hans; Chu, William; Mamedov, Alexandre; Ravi, Ananth; D'Alimonte, Laura; Commisso, Kristina; Helou, Joelle; Deabreu, Andrea; Zhang, Liying; Loblaw, Andrew

    2016-01-01

    Purpose: This study identified predictors of high-grade late hematochezia (HH) following 5-fraction gantry-based stereotactic ablative radiation therapy (SABR). Methods and Materials: Hematochezia data for 258 patients who received 35 to 40 Gy SABR in 5-fractions as part of sequential phase 2 prospective trials was retrieved. Grade 2 or higher late rectal bleeding was labeled HH. Hematochezia needing steroid suppositories, 4% formalin, or 1 to 2 sessions of argon plasma coagulation (APC) was labeled grade 2. More than 2 sessions of APC, blood transfusion, or a course of hyperbaric oxygen was grade 3 and development of visceral fistula, grade 4. Various dosimetric and clinical factors were analyzed using univariate and multivariate analyses. Receiver operating characteristic (ROC) curve analysis and recursive partitioning analysis were used to determine clinically valid cut-off points and identify risk groups, respectively. Results: HH was observed in 19.4%, grade ≥3 toxicity in 3.1%. Median follow-up was 29.7 months (interquartile range [IQR]: 20.6-61.7) Median time to develop HH was 11.7 months (IQR: 9.0-15.2) from the start of radiation. At 2 years, cumulative HH was 4.9%, 27.2%, and 42.1% in patients who received 35 Gy to prostate (4-mm planning target volume [PTV] margin), 40 Gy to prostate (5-mm PTV margin), and 40 Gy to prostate/seminal vesicles (5-mm PTV margin), respectively (P 30%. Conclusions: Rectal V38 and 2 clinical factors were strong predictors of HH following 5-fraction SABR. Planning constraints should keep rectal V38 below 2.0 cm"3.

  13. Surface dimpling on rotating work piece using rotation cutting tool

    Science.gov (United States)

    Bhapkar, Rohit Arun; Larsen, Eric Richard

    2015-03-31

    A combined method of machining and applying a surface texture to a work piece and a tool assembly that is capable of machining and applying a surface texture to a work piece are disclosed. The disclosed method includes machining portions of an outer or inner surface of a work piece. The method also includes rotating the work piece in front of a rotating cutting tool and engaging the outer surface of the work piece with the rotating cutting tool to cut dimples in the outer surface of the work piece. The disclosed tool assembly includes a rotating cutting tool coupled to an end of a rotational machining device, such as a lathe. The same tool assembly can be used to both machine the work piece and apply a surface texture to the work piece without unloading the work piece from the tool assembly.

  14. Earth Rotation

    Science.gov (United States)

    Dickey, Jean O.

    1995-01-01

    The study of the Earth's rotation in space (encompassing Universal Time (UT1), length of day, polar motion, and the phenomena of precession and nutation) addresses the complex nature of Earth orientation changes, the mechanisms of excitation of these changes and their geophysical implications in a broad variety of areas. In the absence of internal sources of energy or interactions with astronomical objects, the Earth would move as a rigid body with its various parts (the crust, mantle, inner and outer cores, atmosphere and oceans) rotating together at a constant fixed rate. In reality, the world is considerably more complicated, as is schematically illustrated. The rotation rate of the Earth's crust is not constant, but exhibits complicated fluctuations in speed amounting to several parts in 10(exp 8) [corresponding to a variation of several milliseconds (ms) in the Length Of the Day (LOD) and about one part in 10(exp 6) in the orientation of the rotation axis relative to the solid Earth's axis of figure (polar motion). These changes occur over a broad spectrum of time scales, ranging from hours to centuries and longer, reflecting the fact that they are produced by a wide variety of geophysical and astronomical processes. Geodetic observations of Earth rotation changes thus provide insights into the geophysical processes illustrated, which are often difficult to obtain by other means. In addition, these measurements are required for engineering purposes. Theoretical studies of Earth rotation variations are based on the application of Euler's dynamical equations to the problem of finding the response of slightly deformable solid Earth to variety of surface and internal stresses.

  15. Toroidal rotation studies in KSTAR

    Science.gov (United States)

    Lee, S. G.; Lee, H. H.; Yoo, J. W.; Kim, Y. S.; Ko, W. H.; Terzolo, L.; Bitter, M.; Hill, K.; KSTAR Team

    2014-10-01

    Investigation of the toroidal rotation is one of the most important topics for the magnetically confined fusion plasma researches since it is essential for the stabilization of resistive wall modes and its shear plays an important role to improve plasma confinement by suppressing turbulent transport. The most advantage of KSTAR tokamak for toroidal rotation studies is that it equips two main diagnostics including the high-resolution X-ray imaging crystal spectrometer (XICS) and charge exchange spectroscopy (CES). Simultaneous core toroidal rotation and ion temperature measurements of different impurity species from the XICS and CES have shown in reasonable agreement with various plasma discharges in KSTAR. It has been observed that the toroidal rotation in KSTAR is faster than that of other tokamak devices with similar machine size and momentum input. This may due to an intrinsically low toroidal field ripple and error field of the KSTAR device. A strong braking of the toroidal rotation by the n = 1 non-resonant magnetic perturbations (NRMPs) also indicates these low toroidal field ripple and error field. Recently, it has been found that n = 2 NRMPs can also damp the toroidal rotation in KSTAR. The detail toroidal rotation studies will be presented. Work supported by the Korea Ministry of Science, ICT and Future Planning under the KSTAR project.

  16. Predictive Parameters of Symptomatic Hematochezia Following 5-Fraction Gantry-Based SABR in Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Musunuru, Hima Bindu [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Davidson, Melanie [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Medical Physics, Odette Cancer Centre, Toronto, Ontario (Canada); Cheung, Patrick; Vesprini, Danny; Liu, Stanley; Chung, Hans; Chu, William [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Mamedov, Alexandre [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Ravi, Ananth [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Medical Physics, Odette Cancer Centre, Toronto, Ontario (Canada); D' Alimonte, Laura [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Commisso, Kristina [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Helou, Joelle [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Deabreu, Andrea [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Zhang, Liying [MacroStat, Inc, Toronto, Ontario (Canada); Loblaw, Andrew, E-mail: andrew.loblaw@sunnybrook.ca [Odette Cancer Centre, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Department of Health Policy, Measurement and Evaluation, University of Toronto, Toronto, Ontario (Canada)

    2016-04-01

    Purpose: This study identified predictors of high-grade late hematochezia (HH) following 5-fraction gantry-based stereotactic ablative radiation therapy (SABR). Methods and Materials: Hematochezia data for 258 patients who received 35 to 40 Gy SABR in 5-fractions as part of sequential phase 2 prospective trials was retrieved. Grade 2 or higher late rectal bleeding was labeled HH. Hematochezia needing steroid suppositories, 4% formalin, or 1 to 2 sessions of argon plasma coagulation (APC) was labeled grade 2. More than 2 sessions of APC, blood transfusion, or a course of hyperbaric oxygen was grade 3 and development of visceral fistula, grade 4. Various dosimetric and clinical factors were analyzed using univariate and multivariate analyses. Receiver operating characteristic (ROC) curve analysis and recursive partitioning analysis were used to determine clinically valid cut-off points and identify risk groups, respectively. Results: HH was observed in 19.4%, grade ≥3 toxicity in 3.1%. Median follow-up was 29.7 months (interquartile range [IQR]: 20.6-61.7) Median time to develop HH was 11.7 months (IQR: 9.0-15.2) from the start of radiation. At 2 years, cumulative HH was 4.9%, 27.2%, and 42.1% in patients who received 35 Gy to prostate (4-mm planning target volume [PTV] margin), 40 Gy to prostate (5-mm PTV margin), and 40 Gy to prostate/seminal vesicles (5-mm PTV margin), respectively (P<.0001). In the ROC analysis, volume of rectum receiving radiation dose of 38 Gy (V38) was a strong predictor of HH with an area under the curve of 0.65. In multivariate analysis, rectal V38 (≥2.0 cm{sup 3}; odds ratio [OR]: 4.7); use of anticoagulants in the follow-up period (OR: 6.5) and presence of hemorrhoids (OR: 2.7) were the strongest predictors. Recursive partitioning analysis showed rectal V38 < 2.0 cm{sup 3}, and use of anticoagulants or rectal V38 ≥ 2.0 cm{sup 3} plus 1 other risk factor resulted in an HH risk of >30%. Conclusions: Rectal V38 and 2

  17. WE-E-18A-11: Fluoro-Tomographic Images From Projections of On-Board Imager (OBI) While Gantry Is Moving

    Energy Technology Data Exchange (ETDEWEB)

    Yi, B; Hu, E; Yu, C; Lasio, G [Univ. of Maryland School Of Medicine, Baltimore, MD (United States)

    2014-06-15

    Purpose: A method to generate a series of fluoro-tomographic images (FTI) of the slice of interest (SOI) from the projection images of the On-board imager (OBI) while gantry is moving is developed and tested. Methods: Tomographic image via background subtraction, TIBS has been published by our group. TIBS uses a priori anatomical information from a previous CT scan to isolate a SOI from a planar kV image by factoring out the attenuations by tissues outside the SOI (background). We extended the idea to 4D TIBS, which enables to generate from the projection of different gantry angles. A set of background images for different angles are prepared. A background image at a given gantry angle is subtracted from the projection image at the same angle to generate a TIBS image. Then the TIBS image is converted to a reference angle. The 4D TIBS is the set of TIBS that originated from gantry angles other than the reference angle. Projection images of lung patients for CBCT acquisition are used to test the 4D TIBS. Results: Fluoroscopic images of a coronal plane of lung patients are acquired from the CBCT projections at different gantry angles and times. Change of morphology of hilar vessels due to breathing and heart beating are visible in the coronal plane, which are generated from the set of the projection images at gantry angles other than antero-posterior. Breathing surrogate or sorting process is not needed. Unlike tomosynthesis, FTI from 4D TIBS maintains the independence of each of the projections thereby reveals temporal variations within the SOI. Conclusion: FTI, fluoroscopic imaging of a SOI with x-ray projections, directly generated from the x-ray projection images at different gantry angles is tested with a lung case and proven feasible. This technique can be used for on-line imaging of moving targets. NIH Grant R01CA133539.

  18. Signaling Device for the Pre-Emergency State of the Elements of the Rotating Assembly of Steam Compressor of Desalination Plant

    Science.gov (United States)

    Danilin, A. I.; Chernyavskiy, A. Zh; Danilin, S. A.; Blagin, E. V.

    2018-01-01

    This article deals with non-contact exploitation control method based on the treatment of the radio wave signal reflected from controlled gear teeth and its advantages in comparison with traditional methods of gear teeth control. Justification of necessity to use such control method during multiplier gears condition determination during its exploitation is given. Also this article deals with influence of different types of gear wear on typical information parameters of analyzed signals. Disadvantages of the method which are the impossibility of determination of certain types of wear are also taken into account. Certain stages of the development of mathematical model for interaction of first converter with controlled surface. Suggested mathematical model uses only the laws of geometric optics without taking wave processes into account but considering first converter direction diagram influence during its interaction with controlled surface. Structural scheme of developed experimental system for gears teeth condition control for steam compressor. Operation of the experimental system of gear control is given on the base of structural scheme. Core of the developed device is microcontroller STM32 which treat the information received from the sensors as well as connection with computer. Certain elements of the experimental control system as well as its components are described separately. Photos of experimental unit for control for control method development in laboratory conditions are presented. Design of the first converter is given in short.

  19. Optimal starting gantry angles using equiangular-spaced beams with intensity modulated radiation therapy for prostate cancer on RTOG 0126: A clinical study of 5 and 7 fields

    International Nuclear Information System (INIS)

    Potrebko, Peter S.; McCurdy, Boyd M.C.; Butler, James B.; El-Gubtan, Adel S.; Nugent, Zoann

    2007-01-01

    Background and Purpose: To investigate the effects of starting gantry angle and number of equiangular-spaced beams for prostate cancer radiotherapy on the Radiation Therapy Oncology Group (RTOG) 0126 protocol using intensity-modulated radiation therapy (IMRT). Materials and methods: Ten localized prostate cancer patients were prescribed to 79.2 Gy in 44 fractions. Static IMRT plans using five and seven equiangular-spaced beams were generated. The starting gantry angles were incremented by 5 o resulting in 15 (5 beams) and 11 (7 beams) plans per patient. Constant target coverage was ensured for all plans in order to isolate the variation in the rectal and bladder metrics as a function of starting gantry angle. Results: The variation with starting gantry angle in rectal metrics using 5 beams was statistically significant (p o and 50 o . Statistically insignificant differences were observed for the bladder metrics using 5 beams. There was little dosimetric variation in the rectal and bladder metrics with 7 beams. Nearly equivalent rectal V 75 Gy was achieved between 5 optimal equiangular-spaced beams starting at 20 o (class solution) and 7 equiangular-spaced beams starting at 0 o for most patients. Conclusions: The use of an optimal starting gantry angle for 5 equiangular-spaced beams, as indicated by a class solution in this study, will facilitate rectal sparing and can produce plans that are equivalent to those employing 7 equiangular-spaced beams

  20. Optimization of multi-slice helical respiration-correlated CT: the effects of table speed and rotation time

    International Nuclear Information System (INIS)

    Wink, Nicole M; McNitt-Gray, Michael F; Solberg, Timothy D

    2005-01-01

    While respiration-correlated CT is gaining acceptance in clinical radiotherapy, the effect of scanning parameters on the image quality has yet to be addressed. The intent of this study was to characterize the effects of gantry rotation and table speed on various image quality characteristics in multi-slice, helical, retrospectively-gated CT images. Images of stationary and moving phantoms were obtained in helical mode on a 20-slice CT scanner. Motion was generated by a computer-controlled platform capable of moving simultaneously in two dimensions. Motion was monitored using a pressure gauge inserted inside an adjustable belt. Selected scans were retrospectively gated into ten phases based on the monitored motion. Gantry rotation speeds of 0.5 s and 1.0 s were evaluated with pitches ranging from 0.1 to 0.45. Several parameters, including calculated object volumes, trajectory (movement from peak to trough), deformation (actual volume divided by volume created with the maximum diameter of contoured object) and z-axis resolution, were used to characterize image quality. These studies indicate that for objects in the peak phase of a movement pattern that simulates breathing, retrospectively gated scans using fast gantry rotation speeds produce volume, trajectory, deformation and z-axis resolution results comparable with those of a stationary object

  1. A study on repainting strategies for treating moderately moving targets with proton pencil beam scanning at the new Gantry 2 at PSI

    International Nuclear Information System (INIS)

    Zenklusen, S M; Pedroni, E; Meer, D

    2010-01-01

    Treating moving targets using a scanning gantry for proton therapy is a promising but very challenging, not yet clinically demonstrated treatment modality. The interference of organ motion with the sequence of the beam delivery produces uncontrolled dose inhomogeneities within the target. One promising approach to overcome this difficulty is to increase the speed of scanning in order to apply the dose repeatedly (so-called repainting). To obtain sufficiently high scanning speeds a new, technologically improved gantry-Gantry 2-has been designed and is currently under construction at PSI. As there are many possible repainting strategies, the way repainting will be implemented on Gantry 2 will depend on the result of a careful analysis of the various treatment delivery strategies available. To achieve this aim, and prior to the start of experimental work with Gantry 2, simulations of dose distribution errors due to organ motion under various beam delivery strategies were investigated. The effects of motion on the dose distribution were studied for moderate motion amplitudes (5 mm) for spherical target volumes in a homogeneous medium and with homogeneous dose. In total over 200 000 dose distributions have been simulated and analyzed and selected results are discussed. From the obtained results we are confident to be able to treat moderately moving targets on Gantry 2 using repainted pencil-beam spot scanning. Continuous line scanning seems to be the most elegant solution; it provides higher repainting rates and produces superior results but is probably more difficult to realize. For larger motion amplitudes, continuous line scanning still shows good results, but we plan anyways to use a gating system for these cases, not only to reduce the inhomogeneity within the target volume but also to reduce safety margins.

  2. NASA safety standard for lifting devices and equipment

    Science.gov (United States)

    1990-09-01

    NASA's minimum safety requirements are established for the design, testing, inspection, maintenance, certification, and use of overhead and gantry cranes (including top running monorail, underhung, and jib cranes), mobile cranes, derrick hoists, and special hoist supported personnel lifting devices (these do not include elevators, ground supported personnel lifts, or powered platforms). Minimum requirements are also addressed for the testing, inspection, and use of Hydra-sets, hooks, and slings. Safety standards are thoroughly detailed.

  3. Strainer device

    International Nuclear Information System (INIS)

    Mokuya, Kenji.

    1975-01-01

    Object: To provide a strainer device, which is adapted to facilitate flushing and is particularly suited for installation in the cooling system of a liquid metal cooled fast breeding reactor. Structure: A casing accommodating a strainer and a blind plate for the selection of a flow path is provided at a suitable portion of the duct line. The blind plate is adapted to be rotated by an opening and closing means consisting of a rod. bellows, shaft and so forth. At the time of flushing, the duct line is sealed by the blind plate. (Nakamura, S.)

  4. Comparative analysis of bacterial decrease by Er:YAG laser radiation and by a tip mounted on a high-rotation device after remove of carious tissue in dentine: study 'in anima mobile'

    International Nuclear Information System (INIS)

    Blay, Claudia Campos

    2001-01-01

    The purpose of this study was to analyze bacterial decrease after removal of carious tissue in dentine by Er:YAG laser, comparing the results with those achieved by using a tip mounted on a high-rotation device. In view of the difficulty of totally eliminating microorganisms from dentine, even after removing all the carious tissue, and since the presence of a smear layer is a proven fact, it would be extremely convenient to eliminate or to achieve better disinfecting of this notoriously contaminated dentine layer, so as to prevent failure of the restoration process in most cases. For this study, we selected 30 pre-molars and molars with caries rated as medium sized and which required restoration treatment. The teeth were divided in two groups, where 15 teeth received conventional treatment (control group - I) and 15 teeth were treated with Er:YAG laser (Kavo Key Laser) (laser group - II). Microbiologic analysis was used in order to determine presence or absence of bacteria. In group I as well as in group II a previous curettage of part of the dentine caries was performed, and the sample was seeded in an Agar culture medium and incubated at 37 deg C for 72 hours, for a quantitative analysis of bacteria colony forming units (c.f.u.) and qualitative bacterial analysis. After this procedure, carious lesions were removed in group I, using a spherical diamond drill mounted on a high rotation tip, followed by cleaning of the cavity with chlorhexidine digluconate at 2%. In group II, an Er:YAG laser emitting a wave length of 2,94 μm, delivering a 250 mJ energy per pulse at a repetition rate of 2 Hz was used for carious tissue removal. Furthermore in group II (laser), the cavity was conditioned by performing a sweeping at all walls of the preparation with the 50/10 fiber (Kavo), in contact mode, using an energy of 60 mJ per pulse at a frequency of 10 Hz. After removing all caries and performing a cleaning operation in both groups, dentine shavings were collected and seeded

  5. Broadband Rotational Spectroscopy

    Science.gov (United States)

    Pate, Brooks

    2014-06-01

    The past decade has seen several major technology advances in electronics operating at microwave frequencies making it possible to develop a new generation of spectrometers for molecular rotational spectroscopy. High-speed digital electronics, both arbitrary waveform generators and digitizers, continue on a Moore's Law-like development cycle that started around 1993 with device bandwidth doubling about every 36 months. These enabling technologies were the key to designing chirped-pulse Fourier transform microwave (CP-FTMW) spectrometers which offer significant sensitivity enhancements for broadband spectrum acquisition in molecular rotational spectroscopy. A special feature of the chirped-pulse spectrometer design is that it is easily implemented at low frequency (below 8 GHz) where Balle-Flygare type spectrometers with Fabry-Perot cavity designs become technologically challenging due to the mirror size requirements. The capabilities of CP-FTMW spectrometers for studies of molecular structure will be illustrated by the collaborative research effort we have been a part of to determine the structures of water clusters - a project which has identified clusters up to the pentadecamer. A second technology trend that impacts molecular rotational spectroscopy is the development of high power, solid state sources in the mm-wave/THz regions. Results from the field of mm-wave chirped-pulse Fourier transform spectroscopy will be described with an emphasis on new problems in chemical dynamics and analytical chemistry that these methods can tackle. The third (and potentially most important) technological trend is the reduction of microwave components to chip level using monolithic microwave integrated circuits (MMIC) - a technology driven by an enormous mass market in communications. Some recent advances in rotational spectrometer designs that incorporate low-cost components will be highlighted. The challenge to the high-resolution spectroscopy community - as posed by Frank De

  6. The first human patient treatment on the PSI spot scanning gantry

    International Nuclear Information System (INIS)

    Munkel, G.; Bevan, A.; Blattmann, H.; Boehringer, T.; Coray, A.; Egger, E.; Lin, S.; Lomax, A.; Pedroni, E.; Schaffner, B.

    1997-01-01

    In November 1996, the first human patient was irradiated with protons on the new PSI facility. The technical and physical reliability of the device had been reviewed and tested extensively; in addition, several animal patients with spontaneous tumors had been treated before and followed up carefully in order to have a biological control and an early warning system for unexpected radiation effects. The first human patient received a boost therapy for a brain metastasis of a malignant melanoma. (author)

  7. Influence of increment of gantry angle and number of arcs on esophageal volumetric modulated arc therapy planning in Monaco planning system: A planning study

    Directory of Open Access Journals (Sweden)

    L Nithya

    2014-01-01

    Full Text Available The objective of this study was to analyze the influence of the increment of gantry angle and the number of arcs on esophageal volumetric modulated arc therapy plan. All plans were done in Monaco planning system for Elekta Synergy linear accelerator with 80 multileaf collimator (MLC. Volumetric modulated arc therapy (VMAT plans were done with different increment of gantry angle like 15 o , 20 o , 30 o and 40 o . The remaining parameters were similar for all the plans. The results were compared. To compare the plan quality with number of arcs, VMAT plans were done with single and dual arc with increment of gantry angle of 20 o . The dose to gross tumor volume (GTV for 60 Gy and planning target volume (PTV for 48 Gy was compared. The dosimetric parameters D 98% , D 95% , D 50% and D max of GTV were analyzed. The homogeneity index (HI and conformity index (CI of GTV were studied and the dose to 98% and 95% of PTV was analyzed. Maximum dose to spinal cord and planning risk volume of cord (PRV cord was compared. The Volume of lung receiving 10 Gy, 20 Gy and mean dose was analyzed. The volume of heart receiving 30 Gy and 45 Gy was compared. The volume of normal tissue receiving greater than 2 Gy and 5 Gy was compared. The number of monitor units (MU required to deliver the plans were compared. The plan with larger increment of gantry angle proved to be superior to smaller increment of gantry angle plans in terms of dose coverage, HI, CI and normal tissue sparing. The number of arcs did not make any difference in the quality of the plan.

  8. Gantry and isocenter displacements of a linear accelerator caused by an add-on micromultileaf collimator

    DEFF Research Database (Denmark)

    Riis, H. L.; Zimmermann, S. J.; Hjelm-Hansen, M.

    2013-01-01

    Purpose: The delivery of high quality stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT) treatments to the patient requires knowledge of the position of the isocenter to submillimeter accuracy. To meet the requirements the deviation between the radiation and mechanical isocenters...... isocenter was measured using a ball-bearing phantom and the electronic portal image device system. These measurements were based on 8 MV photon beams irradiated onto the ball from the four cardinal angles and two opposed collimator angles. The measurements and analysis of the data were carried out...

  9. [Design of SCM inoculation device].

    Science.gov (United States)

    Qian, Mingli; Xie, Haiyuan

    2014-01-01

    The first step of bacilli culture is inoculation bacteria on culture medium. Designing a device to increase efficiency of inoculation is significative. The new device is controlled by SCM. The stepper motor can drive the culture medium rotating, accelerating, decelerating, overturn and suspending. The device is high practicability and efficient, let inoculation easy for operator.

  10. Rotator cuff exercises

    Science.gov (United States)

    ... 25560729 . Read More Frozen shoulder Rotator cuff problems Rotator cuff repair Shoulder arthroscopy Shoulder CT scan Shoulder MRI scan Shoulder pain Patient Instructions Rotator cuff - self-care Shoulder surgery - discharge Using your ...

  11. Shroud brushing device

    International Nuclear Information System (INIS)

    Sakamaki, Kazuo; Takabayashi, Jun-ichi.

    1996-01-01

    A roller screw is assembled at an axial center of the inside of a device main body. A nut is assembled to the axial center of the roller screw secured so as not to be rotated, and a base is integrally connected to the nut. An air cylinder is pivoted to the base by a pin, and an arm is disposed to a piston rod of the air cylinder. The top end of the arm is secured to a motor case, and a brush is rotated by an electric motor in the motor case. The device main body is suspended into the reactor pressure vessel of a BWR type reactor, and seated on the upper surface of a fuel support metal fitting. The roller screw is rotated to move vertically the arm by way of the nut and the base. The device main body is rotated to rotate the arm. The arm is extended and then the electric motor is driven to rotate the brush thereby polishing the inner surface of the shroud. (I.N.)

  12. Shroud brushing device

    Energy Technology Data Exchange (ETDEWEB)

    Sakamaki, Kazuo; Takabayashi, Jun-ichi

    1996-07-12

    A roller screw is assembled at an axial center of the inside of a device main body. A nut is assembled to the axial center of the roller screw secured so as not to be rotated, and a base is integrally connected to the nut. An air cylinder is pivoted to the base by a pin, and an arm is disposed to a piston rod of the air cylinder. The top end of the arm is secured to a motor case, and a brush is rotated by an electric motor in the motor case. The device main body is suspended into the reactor pressure vessel of a BWR type reactor, and seated on the upper surface of a fuel support metal fitting. The roller screw is rotated to move vertically the arm by way of the nut and the base. The device main body is rotated to rotate the arm. The arm is extended and then the electric motor is driven to rotate the brush thereby polishing the inner surface of the shroud. (I.N.)

  13. Stud manipulating device

    International Nuclear Information System (INIS)

    Bunyan, T.W.

    1980-01-01

    A device for inserting and removing studs from bores in a workpiece, for example a nuclear reactor vessel, comprises manipulating devices for operating on individual studs, each capable of tensioning a stud slackening a working nut on the stud, and subsequently removing the stud from the bore. A ring has dogs which can engage working nut recesses to interlock with the nut against relative rotation. Motors coupled to the ring rotate the working nut. A top nut is coupled to the motors to rotate the nut and screw it onto the stud. The top nut with other device parts can be raised and lowered on a tube by a hydraulic actuator. A hydraulic load cell between the top nut and a stool on the workpiece is pressurised to tension the stud by means of the top nut and thus facilitate rotation of the working nut when tightening or slackening. A dog clutch mechanism engages a stud end fitting against relative axial and rotational movement. The mechanism is raised and lowered on a guide member by an actuator. The mechanism has a tubular member and the drive coupling for the motors to the top nut includes a tubular member. Tubular members carry teeth which are engaged when the top nut is raised and the clutch mechanism is lowered, to provide a coupling between the motors and the mechanism for rotating the stud. (U.K.)

  14. Motion as perturbation. II. Development of the method for dosimetric analysis of motion effects with fixed-gantry IMRT

    Energy Technology Data Exchange (ETDEWEB)

    Nelms, Benjamin E. [Canis Lupus LLC, Merrimac, Wisconsin 53561 (United States); Opp, Daniel; Zhang, Geoffrey; Moros, Eduardo; Feygelman, Vladimir, E-mail: vladimir.feygelman@moffitt.org [Department of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612 (United States)

    2014-06-15

    Purpose: In this work, the feasibility of implementing a motion-perturbation approach to accurately estimate volumetric dose in the presence of organ motion—previously demonstrated for VMAT-–is studied for static gantry IMRT. The method's accuracy is improved for the voxels that have very low planned dose but acquire appreciable dose due to motion. The study describes the modified algorithm and its experimental validation and provides an example of a clinical application. Methods: A contoured region-of-interest is propagated according to the predefined motion kernel throughout time-resolved 4D phantom dose grids. This timed series of 3D dose grids is produced by the measurement-guided dose reconstruction algorithm, based on an irradiation of a staticARCCHECK (AC) helical dosimeter array (Sun Nuclear Corp., Melbourne, FL). Each moving voxel collects dose over the dynamic simulation. The difference in dose-to-moving voxel vs dose-to-static voxel in-phantom forms the basis of a motion perturbation correction that is applied to the corresponding voxel in the patient dataset. A new method to synchronize the accelerator and dosimeter clocks, applicable to fixed-gantry IMRT, was developed. Refinements to the algorithm account for the excursion of low dose voxels into high dose regions, causing appreciable dose increase due to motion (LDVE correction). For experimental validation, four plans using TG-119 structure sets and objectives were produced using segmented IMRT direct machine parameters optimization in Pinnacle treatment planning system (v. 9.6, Philips Radiation Oncology Systems, Fitchburg, WI). All beams were delivered with the gantry angle of 0°. Each beam was delivered three times: (1) to the static AC centered on the room lasers; (2) to a static phantom containing a MAPCHECK2 (MC2) planar diode array dosimeter (Sun Nuclear); and (3) to the moving MC2 phantom. The motion trajectory was an ellipse in the IEC XY plane, with 3 and 1.5 cm axes. The period

  15. Reassessment of the Necessity of the Proton Gantry: Analysis of Beam Orientations From 4332 Treatments at the Massachusetts General Hospital Proton Center Over the Past 10 Years

    Energy Technology Data Exchange (ETDEWEB)

    Yan, Susu, E-mail: syan5@mgh.harvard.edu; Lu, Hsiao-Ming; Flanz, Jay; Adams, Judith; Trofimov, Alexei; Bortfeld, Thomas

    2016-05-01

    Purpose: To retrospectively analyze the beam approaches used in gantry-based proton treatments, and to reassess the practical advantages of the gantry, compared with beam approaches that are achievable without a gantry, in the context of present-day technology. Methods and Materials: We reviewed the proton therapy plans of 4332 patients treated on gantries at our hospital, delivered by the double scattering technique (n=4228) and, more recently, pencil beam scanning (PBS) (n=104). Beam approaches, relative to the patient frame, were analyzed individually to identify cases that could be treated without a gantry. Three treatment configurations were considered, with the patient in lying position, sitting position, or both. The FIXED geometry includes a fixed horizontal portal. The BEND geometry enables a limited vertical inflection of the beam by up to 20°. The MOVE geometry allows for flexibility of the patient head and body setup. Results: The percentage of patients with head and neck tumors that could be treated without a gantry using double scattering was 44% in FIXED, 70% in 20° BEND, and 100% in 90° MOVE. For torso regions, 99% of patients could be treated in 20° BEND. Of 104 PBS treatments, all but 1 could be reproduced with FIXED geometry. The only exception would require a 10° BEND capability. Note here that the PBS treatments were applied to select anatomic sites, including only 2 patients with skull-base tumors. Conclusions: The majority of practical beam approaches can be realized with gantry-less delivery, aided by limited beam bending and patient movements. Practical limitations of the MOVE geometry, and treatments requiring a combination of lying and sitting positions, may lower the percentage of head and neck patients who could be treated without a gantry. Further investigation into planning, immobilization, and imaging is needed to remove the practical limitations and to facilitate proton treatment without a gantry.

  16. Device for removing fur

    International Nuclear Information System (INIS)

    Hanawa, Minoru; Nakagawa, Takao; Sakuma, Toyoo; Yonemura, Eizo.

    1976-01-01

    Purpose: To effectively remove fur adhered to fuel rods and to increase working efficiency without use of a lengthy hose. Constitution: In the fur removing device of the present invention, brushes rotated by gears are provided within a casing so that fur adhered to the fuel rods are removed by the brushes and water is rotatably moved by blades housed therein to outwardly blow fur floating in water by means of a centrifugal force. Then, the fur is filtered by a filter outwardly provided. In this way, the fur may be collected within the device to avoid contamination to others. (Kamimura, M.)

  17. An active magnetic regenerator device

    DEFF Research Database (Denmark)

    2015-01-01

    A rotating active magnetic regenerator (AMR) device comprising two or more regenerator beds, a magnet arrangement and a valve arrangement. The valve arrangement comprises a plurality of valve elements arranged substantially immovably with respect to the regenerator beds along a rotational direction...

  18. Dosimetric comparison of RapidArc with fixed gantry dynamic IMRT for loco-regionally advanced nasopharyngeal carcinoma

    International Nuclear Information System (INIS)

    Wu Hao; Han Shukui; Sun Yan; Jiang Fan

    2010-01-01

    Objective: To compare the dosimetric difference of RapidArc and fixed gantry angle dynamic IMRT (dIMRT) for loco-regionally advanced nasopharyngeal carcinoma. Methods: Ten previously treated patients with loco-regionally advanced nasopharyngeal carcinoma were replanned with RapidArc and dIMRT, respectively. The prescription dose was GTV 70 Gy/33 f and PTV 60 Gy/33 f. All plans met the requirement: 95% of PTV was covered by 60 Gy. Dose-volume histogram data, isodose distribution, monitor units, and treatment time were compared. Results: Dose distribution has no significant difference between the two techniques. RapidArc reduced the dose of the brainstem, mandible, and other normal tissues compared with dIMRT. Mean monitor units were 589.5 and 1381.0 for RapidArc and dIMRT (reduced by 57% relatively). Mean treatment time was 2.33 min and 7.82 min for RapidArc and dIMRT (reduced by 70% relatively). Conclusions: Compared with dIMRT, RapidArc achieves equal target coverage and OAR sparing while using fewer monitor units and less time during radiotherapy for patient with loco-regionally advanced nasopharyngeal carcinoma. (authors)

  19. Dosimetric comparison between RapidArc and fixed gantry intensity modulated radiation therapy in treatment of liver carcinoma

    International Nuclear Information System (INIS)

    Ma Changsheng; Yin Yong; Liu Tonghai; Chen Jinhu; Sun Tao; Lin Xiutong

    2010-01-01

    Objective: To compare the dosimetric difference of RapidArc and fixed gantry IMRT for liver carcinoma. Methods: The CT data of 10 liver cancer patients were used to design 3 groups of treatment plan: IMRT plan, single arc RapidArc plan (RA1), and dual arc RapidArc plan (RA2). The planning target volume (PTV) dosimetric distribution, the organs at risk (OAR) dose, the normal tissue dose, mornitor units (MU) and treatment time were compared. Results: The maximum dose of PTV in RA1 and RA2 plans were lower than that of IMRT (Z=-2.0990, -2.666, P 40 of stomach small bowel than IMRT plan, but higher in mean dose of left kidney (Z=-1.988, -2.191, P 5 , V 10 and 15 of healthy tissue in RapidArc plan groups were higher than those in IMRT plan, while the values of V 20 , V 25 and V 30 of healthy tissue in RapidArc plan groups were than those in IMRT plan. The number of computed MU/fraction of Rapid Arc plan was 40% or 46% of IMRT plan and the treatment time was 30% and 40% of IMRT. Conclusions: RapidArc showed improvements in conformity index and healthy tissue sparing with uncompromised target coverage. RapidArc could lead to the less MU and shorter delivery time compared to IMRT. (authors)

  20. Rotational motion in nuclei

    International Nuclear Information System (INIS)

    Bohr, A.

    1977-01-01

    History is surveyed of the development of the theory of rotational states in nuclei. The situation in the 40's when ideas formed of the collective states of a nucleus is evoked. The general rotation theory and the relation between the single-particle and rotational motion are briefly discussed. Future prospects of the rotation theory development are indicated. (I.W.)

  1. Monte Carlo investigation of collapsed versus rotated IMRT plan verification.

    Science.gov (United States)

    Conneely, Elaine; Alexander, Andrew; Ruo, Russell; Chung, Eunah; Seuntjens, Jan; Foley, Mark J

    2014-05-08

    IMRT QA requires, among other tests, a time-consuming process of measuring the absorbed dose, at least to a point, in a high-dose, low-dose-gradient region. Some clinics use a technique of measuring this dose with all beams delivered at a single gantry angle (collapsed delivery), as opposed to the beams delivered at the planned gantry angle (rotated delivery). We examined, established, and optimized Monte Carlo simulations of the dosimetry for IMRT verification of treatment plans for these two different delivery modes (collapsed versus rotated). The results of the simulations were compared to the treatment planning system dose calculations for the two delivery modes, as well as to measurements taken. This was done in order to investigate the validity of the use of a collapsed delivery technique for IMRT QA. The BEAMnrc, DOSXYZnrc, and egs_chamber codes were utilized for the Monte Carlo simulations along with the MMCTP system. A number of different plan complexity metrics were also used in the analysis of the dose distributions in a bid to qualify why verification in a collapsed delivery may or may not be optimal for IMRT QA. Following the Alfonso et al. formalism, the kfclin,frefQclin,Q correction factor was calculated to correct the deviation of small fields from the reference conditions used for beam calibration. We report on the results obtained for a cohort of 20 patients. The plan complexity was investigated for each plan using the complexity metrics of homogeneity index, conformity index, modulation complexity score, and the fraction of beams from a particular plan that intersect the chamber when performing the QA. Rotated QA gives more consistent results than the collapsed QA technique. The kfclin,frefQclin,Qfactor deviates less from 1 for rotated QA than for collapsed QA. If the homogeneity index is less than 0.05 then the kfclin,frefQclin,Q factor does not deviate from unity by more than 1%. A value this low for the homogeneity index can only be obtained

  2. Rotational motion in nuclei

    International Nuclear Information System (INIS)

    Bohr, A.

    1976-01-01

    Nuclear structure theories are reviewed concerned with nuclei rotational motion. The development of the deformed nucleus model facilitated a discovery of rotational spectra of nuclei. Comprehensive verification of the rotational scheme and a successful classification of corresponding spectra stimulated investigations of the rotational movement dynamics. Values of nuclear moments of inertia proved to fall between two marginal values corresponding to rotation of a solid and hydrodynamic pattern of an unrotating flow, respectively. The discovery of governing role of the deformation and a degree of a symmetry violence for determining rotational degrees of freedon is pointed out to pave the way for generalization of the rotational spectra

  3. New methods for optical distance indicator and gantry angle quality control tests in medical linear accelerators: image processing by using a 3D phantom

    Energy Technology Data Exchange (ETDEWEB)

    Shandiz, Mahdi Heravian; Khalilzadeh, Mohammadmahdi; Anvari, Kazem [Mashhad Branch, Islamic Azad University, Mashhad (Iran, Islamic Republic of); Layen, Ghorban Safaeian [Mashhad University of Medical Science, Mashhad (Iran, Islamic Republic of)

    2015-03-15

    In order to keep the acceptable level of the radiation oncology linear accelerators, it is necessary to apply a reliable quality assurance (QA) program. The QA protocols, published by authoritative organizations, such as the American Association of Physicists in Medicine (AAPM), determine the quality control (QC) tests which should be performed on the medical linear accelerators and the threshold levels for each test. The purpose of this study is to increase the accuracy and precision of the selected QC tests in order to increase the quality of treatment and also increase the speed of the tests to convince the crowded centers to start a reliable QA program. A new method has been developed for two of the QC tests; optical distance indicator (ODI) QC test as a daily test and gantry angle QC test as a monthly test. This method uses an image processing approach utilizing the snapshots taken by the CCD camera to measure the source to surface distance (SSD) and gantry angle. The new method of ODI QC test has an accuracy of 99.95% with a standard deviation of 0.061 cm and the new method for gantry angle QC has a precision of 0.43 degrees. The automated proposed method which is used for both ODI and gantry angle QC tests, contains highly accurate and precise results which are objective and the human-caused errors have no effect on the results. The results show that they are in the acceptable range for both of the QC tests, according to AAPM task group 142.

  4. New methods for optical distance indicator and gantry angle quality control tests in medical linear accelerators: image processing by using a 3D phantom

    International Nuclear Information System (INIS)

    Shandiz, Mahdi Heravian; Khalilzadeh, Mohammadmahdi; Anvari, Kazem; Layen, Ghorban Safaeian

    2015-01-01

    In order to keep the acceptable level of the radiation oncology linear accelerators, it is necessary to apply a reliable quality assurance (QA) program. The QA protocols, published by authoritative organizations, such as the American Association of Physicists in Medicine (AAPM), determine the quality control (QC) tests which should be performed on the medical linear accelerators and the threshold levels for each test. The purpose of this study is to increase the accuracy and precision of the selected QC tests in order to increase the quality of treatment and also increase the speed of the tests to convince the crowded centers to start a reliable QA program. A new method has been developed for two of the QC tests; optical distance indicator (ODI) QC test as a daily test and gantry angle QC test as a monthly test. This method uses an image processing approach utilizing the snapshots taken by the CCD camera to measure the source to surface distance (SSD) and gantry angle. The new method of ODI QC test has an accuracy of 99.95% with a standard deviation of 0.061 cm and the new method for gantry angle QC has a precision of 0.43 degrees. The automated proposed method which is used for both ODI and gantry angle QC tests, contains highly accurate and precise results which are objective and the human-caused errors have no effect on the results. The results show that they are in the acceptable range for both of the QC tests, according to AAPM task group 142.

  5. Highest performance in 3D metal cutting at smallest footprint: benchmark of a robot based system vs. parameters of gantry systems

    Science.gov (United States)

    Scheller, Torsten; Bastick, André; Michel-Triller, Robert; Manzella, Christon

    2014-02-01

    In the automotive industry as well as in other industries ecological aspects regarding energy savings are driving new technologies and materials, e.g. lightweight materials as aluminium or press hardened steels. Processing such parts especially complex 3D shaped parts laser manufacturing has become the key process offering highest efficiency. The most established systems for 3D cutting applications are based on gantry systems. The disadvantage of those systems is their huge footprint to realize the required stability and work envelope. Alternatively a robot based system might be of advantage if accuracy, speed and overall performance would be capable processing automotive parts. With the BIM "beam in motion" system, JENOPTIK Automatisierungstechnik GmbH has developed a modular robot based laser processing machine, which meets all OEM specs processing press hardened steel parts. A benchmark of the BIM versus a gantry system was done regarding all required parameters to fulfil OEM specifications for press hardened steel parts. As a result a highly productive, accurate and efficient system can be described based on one or multiple robot modules working simultaneously together. The paper presents the improvements on the robot machine concept BIM addressed in 2012 [1] leading to an industrial proven system approach for the automotive industry. It further compares the performance and the parameters for 3D cutting applications of the BIM system versus a gantry system by samples of applied parts. Finally an overview of suitable applications for processing complex 3D parts with high productivity at small footprint is given.

  6. SU-E-T-512: Evaluation of Treatment Planning Dose Calculation Accuracy at the Interface of Prosthetic Devices.

    Science.gov (United States)

    Paulu, D; Alaei, P

    2012-06-01

    To evaluate the ability of treatment planning algorithm to accurately predict dose delivered at the interface of high density implanted devices. A high density (7.6 g/cc) Cobalt-Chromium-Molybdenum hip prosthesis was molded into an epoxy-based cylindrical leg phantom. The phantom was designed to be separated in half to access the prosthesis and to place the TLDs. Using MVCT to image the apparatus, a simple treatment plan was developed using the Philips Pinnacle treatment planning system. Wires were placed in the molded epoxy to allow for accurate definition of measurement sites (TLD positions) along the surface of the prosthesis. Micro-cube TLDs (1 mm 3 ) were placed at six measurement locations for which the dose had been calculated by the treatment planning system. An Elekta Synergy linear accelerator was used to deliver a 400 cGy plan to the phantom with 6 MV photons in a single fraction. A total of four 10 cm × 21 cm fields were used at 0, 90, 180, and 270 degree gantry rotations. Initial results indicate that the measured dose is 7-17% lower than the dose calculated by the treatment planning system. Further study using high energy beams are also in progress. Initial results indicate that the treatment planning system does predict the dose near a high density prosthetic device within 10-15% but underestimates the dose. The results of this study could help in designing treatment plans which would reduce the uncertainty of the dose delivered in the vicinity of prosthetic hip implants and similar devices. © 2012 American Association of Physicists in Medicine.

  7. Development of NTD Hydraulic Rotation System for Kijang Research Reactor

    International Nuclear Information System (INIS)

    Kang, Hanok; Park, Kijung; Park, Yongsoo; Kim, Seong Hoon; Park, Cheol

    2014-01-01

    The KJRR will be mainly utilized for isotope production, NTD (Neutron Transmutation Doping) production, and related research activities. During irradiation for the NTD process, the irradiation rigs containing the silicon ingot rotate at a constant speed to ensure precisely defined homogeneity of the irradiation. The NTDHRS requires only hydraulic piping conveniently routed to the rotating devices inside the reactor pool. The resulting layout leaves the pool area clear of obstructions which might obscure vision and hinder target handling for operators. Pump banks and control valves are located remotely in a dedicated plant room allowing easy access and online maintenance. The necessities and major characteristic of NTD hydraulic rotation system are described in this study. A new NTD hydraulic rotation system are being developed to rotate the irradiation rigs at a constant speed and supply cooling flow for the irradiation rigs and reflector assembly. The configuration of the NTD hydraulic rotation device is discussed and practical methods to improve the rotational performance are suggested

  8. Development of NTD Hydraulic Rotation System for Kijang Research Reactor

    Energy Technology Data Exchange (ETDEWEB)

    Kang, Hanok; Park, Kijung; Park, Yongsoo; Kim, Seong Hoon; Park, Cheol [Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)

    2014-05-15

    The KJRR will be mainly utilized for isotope production, NTD (Neutron Transmutation Doping) production, and related research activities. During irradiation for the NTD process, the irradiation rigs containing the silicon ingot rotate at a constant speed to ensure precisely defined homogeneity of the irradiation. The NTDHRS requires only hydraulic piping conveniently routed to the rotating devices inside the reactor pool. The resulting layout leaves the pool area clear of obstructions which might obscure vision and hinder target handling for operators. Pump banks and control valves are located remotely in a dedicated plant room allowing easy access and online maintenance. The necessities and major characteristic of NTD hydraulic rotation system are described in this study. A new NTD hydraulic rotation system are being developed to rotate the irradiation rigs at a constant speed and supply cooling flow for the irradiation rigs and reflector assembly. The configuration of the NTD hydraulic rotation device is discussed and practical methods to improve the rotational performance are suggested.

  9. Rotating Reverse-Osmosis for Water Purification

    Science.gov (United States)

    Lueptow, RIchard M.

    2004-01-01

    A new design for a water-filtering device combines rotating filtration with reverse osmosis to create a rotating reverse- osmosis system. Rotating filtration has been used for separating plasma from whole blood, while reverse osmosis has been used in purification of water and in some chemical processes. Reverse- osmosis membranes are vulnerable to concentration polarization a type of fouling in which the chemicals meant not to pass through the reverse-osmosis membranes accumulate very near the surfaces of the membranes. The combination of rotating filtration and reverse osmosis is intended to prevent concentration polarization and thereby increase the desired flux of filtered water while decreasing the likelihood of passage of undesired chemical species through the filter. Devices based on this concept could be useful in a variety of commercial applications, including purification and desalination of drinking water, purification of pharmaceutical process water, treatment of household and industrial wastewater, and treatment of industrial process water. A rotating filter consists of a cylindrical porous microfilter rotating within a stationary concentric cylindrical outer shell (see figure). The aqueous suspension enters one end of the annulus between the inner and outer cylinders. Filtrate passes through the rotating cylindrical microfilter and is removed via a hollow shaft. The concentrated suspension is removed at the end of the annulus opposite the end where the suspension entered.

  10. Rotary combustion device

    NARCIS (Netherlands)

    2008-01-01

    Rotary combustion device (1) with rotary combustion chamber (4). Specific measures are taken to provide ignition of a combustible mixture. It is proposed that a hollow tube be provided coaxially with the axis of rotation (6), so that a small part of the mixture is guided into the combustion chamber.

  11. Radiation area monitor device and method

    Science.gov (United States)

    Vencelj, Matjaz; Stowe, Ashley C.; Petrovic, Toni; Morrell, Jonathan S.; Kosicek, Andrej

    2018-01-30

    A radiation area monitor device/method, utilizing: a radiation sensor; a rotating radiation shield disposed about the radiation sensor, wherein the rotating radiation shield defines one or more ports that are transparent to radiation; and a processor operable for analyzing and storing a radiation fingerprint acquired by the radiation sensor as the rotating radiation shield is rotated about the radiation sensor. Optionally, the radiation sensor includes a gamma and/or neutron radiation sensor. The device/method selectively operates in: a first supervised mode during which a baseline radiation fingerprint is acquired by the radiation sensor as the rotating radiation shield is rotated about the radiation sensor; and a second unsupervised mode during which a subsequent radiation fingerprint is acquired by the radiation sensor as the rotating radiation shield is rotated about the radiation sensor, wherein the subsequent radiation fingerprint is compared to the baseline radiation fingerprint and, if a predetermined difference threshold is exceeded, an alert is issued.

  12. Rotationally invariant correlation filtering

    International Nuclear Information System (INIS)

    Schils, G.F.; Sweeney, D.W.

    1985-01-01

    A method is presented for analyzing and designing optical correlation filters that have tailored rotational invariance properties. The concept of a correlation of an image with a rotation of itself is introduced. A unified theory of rotation-invariant filtering is then formulated. The unified approach describes matched filters (with no rotation invariance) and circular-harmonic filters (with full rotation invariance) as special cases. The continuum of intermediate cases is described in terms of a cyclic convolution operation over angle. The angular filtering approach allows an exact choice for the continuous trade-off between loss of the correlation energy (or specificity regarding the image) and the amount of rotational invariance desired

  13. SU-G-TeP4-03: A Multileaf Collimator Calibration and Quality Assurance Technique Using An Electronic Portal Imaging Device

    Energy Technology Data Exchange (ETDEWEB)

    Lebron, S; Yan, G; Li, J; Lu, B; Liu, C [University of Florida, Gainesville, FL (United States)

    2016-06-15

    Purpose: To develop an accurate and quick multileaf collimator (MLC) calibration and quality assurance technique using an electronic portal imaging device (EPID) Methods: The MLC models used include the MLCi and Agility (Elekta Ltd). This technique consists of two 22(L)x10(W) cm{sup 2} fields with 0{sup 0} and 180{sup 0} collimator angles centered to an offset EPID. The MLC opening is estimated by calculating the profile at the image’s center in the image’s horizontal direction. Scans in the image’s vertical direction were calculated every 20 pixels in the inner 70% of estimated MLC opening. The profiles’ edges were fitted with linear equations to determine the image’s rotation angle. Then, crossline profiles were scanned at the center of each leaf taking into account the leaf’s width at isocenter and the rotation angle. The profiles’ edges determine the location of the leaves’ edges and these were subtracted from the reference leaf’s position in order to determine the relative leaf offsets. The edge location of all profiles was determined by using the parameterized gradient of the penumbra region. The technique was tested against an established diode array-based method, and for different MLC systems, patterns, gantry angles, days, energies, beam modalities and MLC openings. Results: The differences between the proposed and established methods were 0.26±0.19mm. The leaf offsets’ deviation was <0.3mm (5 months period). For pattern fields, the differences between predetermined and calculated offsets were 0.18±0.18mm. The leaf offset deviation of measurements with different energies and MLC openings were <0.1mm and <0.3mm, respectively. The differences between offsets of FF and FFF beams were 0.01±0.02mm (<0.07mm). The differences between the offsets at different gantry angles were 0.08±0.15mm. Conclusion: The proposed method proved to be accurate and efficient in calculating the relative leaf offsets. Parameterized field edge is essential to

  14. A Failure Detection Strategy for Intrafraction Prostate Motion Monitoring With On-Board Imagers for Fixed-Gantry IMRT

    International Nuclear Information System (INIS)

    Liu Wu; Luxton, Gary; Xing Lei

    2010-01-01

    Purpose: To develop methods to monitor prostate intrafraction motion during fixed-gantry intensity-modulated radiotherapy using MV treatment beam imaging together with minimal kV imaging for a failure detection strategy that ensures prompt detection when target displacement exceeds a preset threshold. Methods and Materials: Real-time two-dimensional (2D) marker position in the MV image plane was obtained by analyzing cine-MV images. The marker's in-line movement, and thus its time-varying three-dimensional (3D) position, was estimated by combining the 2D projection data with a previously established correlative relationship between the directional components of prostate motion. A confirmation request for more accurate localization using MV-kV triangulation was triggered when the estimated prostate displacement based on the cine-MV data was greater than 3 mm. An interventional action alert followed on positive MV-kV confirmation. To demonstrate the feasibility and accuracy of the proposed method, simulation studies of conventional-fraction intensity-modulated radiotherapy sessions were done using 536 Calypso-measured prostate trajectories from 17 radiotherapy patients. Results: A technique for intrafraction prostate motion management has been developed. The technique, using 'freely available' cine-MV images and minimum on-board kV imaging (on average 2.5 images/fraction), successfully limited 3D prostate movement to within a range of 3 mm relative to the MV beam for 99.4% of the total treatment time. On average, only approximately one intervention/fraction was needed to achieve this level of accuracy. Conclusion: Instead of seeking to accurately and continuously localize the prostate target as existing motion tracking systems do, the present technique effectively uses cine-MV data to provide a clinically valuable way to minimize kV usage, while maintaining high targeting accuracy.

  15. Diagnostic accuracy of state-of-the-art MDCT scanners without gantry tilt in patients with oral and oropharyngeal cancer

    International Nuclear Information System (INIS)

    Bannas, Peter; Habermann, Christian R.; Jung, Caroline; Bley, Thorsten A.; Ittrich, Harald; Adam, Gerhard; Koops, Andreas

    2012-01-01

    Purpose: Current CT-protocols for staging oral and oropharyngeal cancer include primary transversal slices and secondary tilted slices to avoid artifact-producing regions of dental metalwork. Some of the latest MDCT scanners do not allow gantry tilt. Hence, we assessed the relevance of secondary tilted slices in tumor staging. Materials and methods: Scans of a tiltable 64-row MDCT-scanner of 82 patients with oral or oropharyngeal cancer were retrospectively and independently evaluated twice by three readers: once using the primary transversal scans only, and once taking the transversal and secondary tilted scans into account. Tumor diameters and TN-stage were determined with both methods. Artifacts on transversal scans were analyzed using a 3-point-ranking-scale. Results: Image quality was impaired by severe artifacts in 24% of transversal slices of the oral cavity and in 12% of the oropharynx. The three readers detected an average of 57.7 ± 2.1 of 82 tumors (70%) on transversal CT slices. An average of 6.3 ± 0.6 more tumors (8%) were detected when transversal studies were evaluated in conjunction with secondary tilted slices, leading to a significantly (p = 0.0156–0.0313) increased average detection rate of 64.0 ± 2.0 tumors (78%). Moreover, secondary tilted slices led to a correction of underestimated tumor stages in up to six patients (7.3%). Conclusion: Tilted slices that avoid artifact-producing regions of dental metalwork significantly improve the reader's sensitivity and are of incremental value for staging of oral and oropharyngeal cancers.

  16. Inducer Hydrodynamic Load Measurement Devices

    Science.gov (United States)

    Skelley, Stephen E.; Zoladz, Thomas F.

    2002-01-01

    Marshall Space Flight Center (MSFC) has demonstrated two measurement devices for sensing and resolving the hydrodynamic loads on fluid machinery. The first - a derivative of the six component wind tunnel balance - senses the forces and moments on the rotating device through a weakened shaft section instrumented with a series of strain gauges. This "rotating balance" was designed to directly measure the steady and unsteady hydrodynamic loads on an inducer, thereby defining both the amplitude and frequency content associated with operating in various cavitation modes. The second device - a high frequency response pressure transducer surface mounted on a rotating component - was merely an extension of existing technology for application in water. MSFC has recently completed experimental evaluations of both the rotating balance and surface-mount transducers in a water test loop. The measurement bandwidth of the rotating balance was severely limited by the relative flexibility of the device itself, resulting in an unexpectedly low structural bending mode and invalidating the higher frequency response data. Despite these limitations, measurements confirmed that the integrated loads on the four-bladed inducer respond to both cavitation intensity and cavitation phenomena. Likewise, the surface-mount pressure transducers were subjected to a range of temperatures and flow conditions in a non-rotating environment to record bias shifts and transfer functions between the transducers and a reference device. The pressure transducer static performance was within manufacturer's specifications and dynamic response accurately followed that of the reference.

  17. Fundamentals of semiconductor devices

    CERN Document Server

    Lindmayer, Joseph

    1965-01-01

    Semiconductor properties ; semiconductor junctions or diodes ; transistor fundamentals ; inhomogeneous impurity distributions, drift or graded-base transistors ; high-frequency properties of transistors ; band structure of semiconductors ; high current densities and mechanisms of carrier transport ; transistor transient response and recombination processes ; surfaces, field-effect transistors, and composite junctions ; additional semiconductor characteristics ; additional semiconductor devices and microcircuits ; more metal, insulator, and semiconductor combinations for devices ; four-pole parameters and configuration rotation ; four-poles of combined networks and devices ; equivalent circuits ; the error function and its properties ; Fermi-Dirac statistics ; useful physical constants.

  18. A planning and delivery study of a rotational IMRT technique with burst delivery

    International Nuclear Information System (INIS)

    Kainz, Kristofer; Chen, Guang-Pei; Chang, Yu-Wen; Prah, Douglas; Sharon Qi, X.; Shukla, Himanshu P.; Stahl, Johannes; Allen Li, X.

    2011-01-01

    Purpose: A novel rotational IMRT (rIMRT) technique using burst delivery (continuous gantry rotation with beam off during MLC repositioning) is investigated. The authors evaluate the plan quality and delivery efficiency and accuracy of this dynamic technique with a conventional flat 6 MV photon beam. Methods: Burst-delivery rIMRT was implemented in a planning system and delivered with a 160-MLC linac. Ten rIMRT plans were generated for five anonymized patient cases encompassing head and neck, brain, prostate, and prone breast. All plans were analyzed retrospectively and not used for treatment. Among the varied plan parameters were the number of optimization points, number of arcs, gantry speed, and gantry angle range (alpha) over which the beam is turned on at each optimization point. Combined rotational/step-and-shoot rIMRT plans were also created by superimposing multiple-segment static fields at several optimization points. The rIMRT trial plans were compared with each other and with plans generated using helical tomotherapy and VMAT. Burst-mode rotational IMRT plans were delivered and verified using a diode array, ionization chambers, thermoluminescent dosimeters, and film. Results: Burst-mode rIMRT can achieve plan quality comparable to helical tomotherapy, while the former may lead to slightly better OAR sparing for certain cases and the latter generally achieves slightly lower hot spots. Few instances were found in which increasing the number of optimization points above 36, or superimposing step-and-shoot IMRT segments, led to statistically significant improvements in OAR sparing. Using an additional rIMRT partial arc yielded substantial OAR dose improvements for the brain case. Measured doses from the rIMRT plan delivery were within 4% of the plan calculation in low dose gradient regions. Delivery time range was 228-375 s for single-arc rIMRT 200-cGy prescription with a 300 MU/min dose rate, comparable to tomotherapy and VMAT. Conclusions: Rotational IMRT

  19. Parameterization of rotational spectra

    International Nuclear Information System (INIS)

    Zhou Chunmei; Liu Tong

    1992-01-01

    The rotational spectra of the strongly deformed nuclei with low rotational frequencies and weak band mixture are analyzed. The strongly deformed nuclei are commonly encountered in the rare-earth region (e. g., 150 220). A lot of rotational band knowledge are presented

  20. Rotating reactors : a review

    NARCIS (Netherlands)

    Visscher, F.; Schaaf, van der J.; Nijhuis, T.A.; Schouten, J.C.

    2013-01-01

    This review-perspective paper describes the current state-of-the-art in the field of rotating reactors. The paper has a focus on rotating reactor technology with applications at lab scale, pilot scale and industrial scale. Rotating reactors are classified and discussed according to their geometry:

  1. Rotating Stars in Relativity

    Directory of Open Access Journals (Sweden)

    Stergioulas Nikolaos

    2003-01-01

    Full Text Available Rotating relativistic stars have been studied extensively in recent years, both theoretically and observationally, because of the information they might yield about the equation of state of matter at extremely high densities and because they are considered to be promising sources of gravitational waves. The latest theoretical understanding of rotating stars in relativity is reviewed in this updated article. The sections on the equilibrium properties and on the nonaxisymmetric instabilities in f-modes and r-modes have been updated and several new sections have been added on analytic solutions for the exterior spacetime, rotating stars in LMXBs, rotating strange stars, and on rotating stars in numerical relativity.

  2. On Averaging Rotations

    DEFF Research Database (Denmark)

    Gramkow, Claus

    1999-01-01

    In this article two common approaches to averaging rotations are compared to a more advanced approach based on a Riemannian metric. Very offten the barycenter of the quaternions or matrices that represent the rotations are used as an estimate of the mean. These methods neglect that rotations belo...... approximations to the Riemannian metric, and that the subsequent corrections are inherient in the least squares estimation. Keywords: averaging rotations, Riemannian metric, matrix, quaternion......In this article two common approaches to averaging rotations are compared to a more advanced approach based on a Riemannian metric. Very offten the barycenter of the quaternions or matrices that represent the rotations are used as an estimate of the mean. These methods neglect that rotations belong...

  3. A new device of CT-targeted percutaneous punctures

    International Nuclear Information System (INIS)

    Ozdoba, C.

    1990-01-01

    This paper reports on a device for CT-targeted percutaneous punctures. After extensive phantom studies, punctures of vertebral disk space (diskography/nucleotomy) and orbit retrobulbar anesthesia/biopsy were performed in 15 patients. The new device is easy to mount on the CT scanner and does not require additional devices affixed to the patient. Its design and construction permit exact calculation of needle angle and depth, even when the gantry is tilted. Percutaneous biopsies can be performed in approximately 40 minutes. The device is accurate within ±0.5 mm, thereby allowing puncture of even small lesions deep in the orbital cone. Retrobulbar anesthesia can be performed safely in patients with a long myopic bulbus; in nucleotomy, instant control of the instrument's position within the disk space is possible

  4. Nuclear fuel powder transfer device

    International Nuclear Information System (INIS)

    Komono, Akira

    1998-01-01

    A pair of parallel rails are laid between a receiving portion to a molding portion of a nuclear fuel powder transfer device. The rails are disposed to the upper portion of a plurality of parallel support columns at the same height. A powder container is disposed while being tilted in the inside of the vessel main body of a transfer device, and rotational shafts equipped with wheels are secured to right and left external walls. A nuclear powder to be mixed, together with additives, is supplied to the powder container of the transfer device. The transfer device engaged with the rails on the receiving side is transferred toward the molding portion. The wheels are rotated along the rails, and the rotational shafts, the vessel main body and the powder container are rotated. The nuclear powder in the tilted powder container disposed is rotated right and left and up and down by the rotation, and the powder is mixed satisfactory when it reaches the molding portion. (I.N.)

  5. Freely tunable broadband polarization rotator for terahertz waves.

    Science.gov (United States)

    Fan, Ren-Hao; Zhou, Yu; Ren, Xiao-Ping; Peng, Ru-Wen; Jiang, Shang-Chi; Xu, Di-Hu; Xiong, Xiang; Huang, Xian-Rong; Wang, Mu

    2015-02-18

    A freely tunable polarization rotator for broadband terahertz waves is demonstrated using a three-rotating-layer metallic grating structure, which can conveniently rotate the polarization of a linearly polarized terahertz wave to any desired direction with nearly perfect conversion efficiency. This low-cost, high-efficiency, and freely tunable device has potential applications as material analysis, wireless communication, and THz imaging. © 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  6. Reversible Polarization Rotation in Epitaxial Ferroelectric Bilayers

    DEFF Research Database (Denmark)

    Liu, Guangqing; Zhang, Qi; Huang, Hsin-Hui

    2016-01-01

    Polarization rotation engineering is a promising path to giant dielectric and electromechanical responses in ferroelectric materials and devices. This work demonstrates robust and reversible in- to out-of-plane polarization rotation in ultrathin (nanoscale) epitaxial (001) tetragonal PbZr0.3Ti0.7O3...... large-scale polarization rotation switching (≈60 μC cm−2) and an effective d 33 response 500% (≈250 pm V−1) larger than the PZT-R layer alone. Furthermore, this enhancement is stable for more than 107 electrical switching cycles. These bilayers present a simple and highly controllable means to design...... and optimize rotational polar systems as an alternate to traditional composition-based approaches. The precise control of the subtle interface-driven interactions between the lattice and the external factors that control polarization opens a new door to enhanced—or completely new—functional properties....

  7. The complex challenge of repairing the gantry steelwork on the first generation magnox storage pond at Sellafield: Legacy Waste Storage, First Generation Magnox Storage Pond - 59133

    International Nuclear Information System (INIS)

    Richardson, Ian E.

    2012-01-01

    This paper puts into context the challenges that were faced when repairing the Gantry Steelwork of the First Generation Magnox Storage Pond (FGMSP). The First Generation Magnox Fuel Storage Pond (FGMSP) provided fuel storage and de-canning capability from the early 1960's until 1986. A significant programme of work has been underway since the completion of operational activities to support the programmes strategic intent of retrieving and storing all legacy wastes, and remediating the structure of the plant to support decommissioning activities. A key enabler to the retrievals programme is the Pond Skip Handler Machine (SHM), removed from service in 2002 following the discovery of significant signs of corrosion and distress, an inevitable consequence of being located in a coastal, salt laden environment. The SHM provides sole capability to access and retrieve the inventory of over 1000 fuel skips. It is also fundamental to future operations and the deployment of de-sludging equipment to recover significant bulk sludge's from the pond floor. Failure of the SHM steelwork gantry at worst case could potentially result in the Skip Handler Machine being derailed. This has some potential to damage to the pond structure and at worst case may result in local radiological and environmental consequences. This paper will examine the challenges faced by the team as they successfully defined, planned and executed remedial work to a specific aspect of the civil structure, the SHM gantry rail system, using a purpose built refurbishment platform; the Gantry Refurbishment System. The paper will examine how an 'innovative' approach was adopted to resolve the related issues of: - Refurbishing an aged structure to meet extended future operational demands. - The application of pragmatic engineering solutions against current codes and standards including seismic performance; - Provision of safe access for the workforce to undertake the refurbishment work against significant radiological

  8. Determination of the attenuation due to the treatment table according to the angle of gantry for system treatments verification Delta4

    International Nuclear Information System (INIS)

    Linares Doblado, R.; Mateo Rodriguez, B.; Perucha Ortega, M.

    2013-01-01

    The optimal solution to the problem of absorption into the table would be to incorporate it into the system of planning as one structure more, but not all planners incorporate this functionality. As an alternative solution, approximate, this paper evaluates through measures in the system mentioned effect, with the aim of obtaining a table of absorption vs. angle of gantry, for different field sizes that will be used to correct the dose values recorded by the QA system. Research on the effect of the corrections in the histograms of dose and gamma. (Author)

  9. View of the moving head of the gantry machine and the working area containingthe supply and assembly platforms (trays in green).

    CERN Multimedia

    Alan Honma

    1999-01-01

    The robotic assembly machine consists of the gantry positioning systemoutfitted with pickup tooling heads and camera+microscope for accurateposition measurements. The procedure is to place the components onthe working platforms and the machine applies glue, picks and placesthe silicon sensors and front-end hybrids onto the frames. The components are held in place by vacuum to prevent movement untilthe glue has cured. Up to four modules can be assembled at one time.The platforms are removable allowing assembly to continue on a newset of modules.

  10. Computed tomography in endocrine orbitopathy: Effects of different gantry tilt and patient positioning on measurements of eye muscle thickness, and possibilities for correction

    International Nuclear Information System (INIS)

    Markl, A.; Hilbertz, T.; Mayr, B.; Lissner, J.; Pickardt, C.R.

    1986-01-01

    Thickening of eye muscles in endocrine orbitopathy can be demonstrated particularly impressively in coronary computed tomograms. However, when measuring the height and width of rectus eye muscles manifesting pathologic changes, the measurement is increased by a deviation from the coronary section plane; this is due to different tilting of the gantry. This often leads to an incorrect stage classification and makes objective observation of the course (e.g., under therapy) impossible. By converting the measured values into the actual extent of the muscles by means of the cosine set, appreciable changes in the pattern and frequency of affection of the rectus eye muscles were found in 121 patients examined. (orig.) [de

  11. Current Biomechanical Concepts for Rotator Cuff Repair

    Science.gov (United States)

    2013-01-01

    For the past few decades, the repair of rotator cuff tears has evolved significantly with advances in arthroscopy techniques, suture anchors and instrumentation. From the biomechanical perspective, the focus in arthroscopic repair has been on increasing fixation strength and restoration of the footprint contact characteristics to provide early rehabilitation and improve healing. To accomplish these objectives, various repair strategies and construct configurations have been developed for rotator cuff repair with the understanding that many factors contribute to the structural integrity of the repaired construct. These include repaired rotator cuff tendon-footprint motion, increased tendon-footprint contact area and pressure, and tissue quality of tendon and bone. In addition, the healing response may be compromised by intrinsic factors such as decreased vascularity, hypoxia, and fibrocartilaginous changes or aforementioned extrinsic compression factors. Furthermore, it is well documented that torn rotator cuff muscles have a tendency to atrophy and become subject to fatty infiltration which may affect the longevity of the repair. Despite all the aforementioned factors, initial fixation strength is an essential consideration in optimizing rotator cuff repair. Therefore, numerous biomechanical studies have focused on elucidating the strongest devices, knots, and repair configurations to improve contact characteristics for rotator cuff repair. In this review, the biomechanical concepts behind current rotator cuff repair techniques will be reviewed and discussed. PMID:23730471

  12. Gating-by-rotation: a solution to the problem of intratreatment motion in helical tomotherapy

    International Nuclear Information System (INIS)

    Kapatoes, J.M.; Olivera, G.H.; Schloesser, E.A.; Pearson, D.W.; Balog, J.P.; Ruchala, K.J.; Schmidt, R.; Reckwerdt, P.J.; Mehta, M.P.; Mackie, T.R.

    2001-01-01

    Purpose: To assess the feasibility of addressing intratreatment motion issues in helical tomotherapy by gating the treatments by rotation. Intratreatment motion is a problem common to all IMRT techniques. Traditional methods of gating in conformal radiotherapy and some forms of IMRT are not applicable to helical tomotherapy due to the continuous rotation of the gantry. An alternative method is presented. Materials and Methods: Rotation-gating in helical tomotherapy is the process in which one rotation of treatment is immediately followed by a rotation of non-treatment. This on-off strategy is repeated for the full treatment volume. During the treatment rotations, the patient is required to hold their breath while the intensity-modulated fan beam deposits dose. For the non-treatment rotations, the patient is allowed to breathe freely as all leaves of the MLC will be closed, the accelerator disabled, or both. The couch indexes normally for treatment rotations and holds the patient stationary during non-treatment rotations. An investigation was conducted to assess the feasibility of rotation-gating. Film was placed between two hemispheres of a water phantom and a continuous helical delivery was carried out with all leaves opened. The film was replaced and another treatment was performed employing rotation-gating. The two films were compared to assess the process. The films were irradiated to dose levels within the linear region of the film response curve (maximum film dose ∼35 cGy). Films were also acquired with all leaves closed to quantify leakage dose through the collimation systems. Results: Central profiles for the inferior-superior direction (parallel to the direction of translation) for both films are displayed in Figure 1. The profiles agree very well, illustrating that a rotation-gated treatment closely mimics a continuous helical delivery. The only significant discrepancy lay in the tails of the profiles: a higher film dose is seen for the rotation

  13. Reluctance device

    International Nuclear Information System (INIS)

    Claridge, A.N.; Smith, A.J.

    1983-01-01

    A reluctance device comprises two or more phases, each of which has a rotor mounted for rotation within a tubular member and an annular stator positioned externally of the tubular member. The rotor and the internal surface of the tubular member are each provided with aligned, axially spaced apart annular arrays of teeth, the teeth on the rotor confronting those on the tubular member in radially spaced apart relationship. The stator encloses a coil which, when electrically energised, creates a plurality of magnetic flux paths each of which extends radially between the rotor and stator via the confronting teeth and the tubular member, and axially along both the rotor and the portion of the stator located radially outwardly of the coil. The portion of the tubular member intermediate the teeth thereon is provided with a non-magnetic insert in order to resist the axial passage of magnetic flux therethrough. In one of the claims, the tubular member comprises a portion of a nuclear reactor which is adapted to contain a reactor control rod, the rotor constituting a portion of the drive mechanism for the control rod contained in operation within the tubular member. (author)

  14. Rotations with Rodrigues' vector

    International Nuclear Information System (INIS)

    Pina, E

    2011-01-01

    The rotational dynamics was studied from the point of view of Rodrigues' vector. This vector is defined here by its connection with other forms of parametrization of the rotation matrix. The rotation matrix was expressed in terms of this vector. The angular velocity was computed using the components of Rodrigues' vector as coordinates. It appears to be a fundamental matrix that is used to express the components of the angular velocity, the rotation matrix and the angular momentum vector. The Hamiltonian formalism of rotational dynamics in terms of this vector uses the same matrix. The quantization of the rotational dynamics is performed with simple rules if one uses Rodrigues' vector and similar formal expressions for the quantum operators that mimic the Hamiltonian classical dynamics.

  15. On Averaging Rotations

    DEFF Research Database (Denmark)

    Gramkow, Claus

    2001-01-01

    In this paper two common approaches to averaging rotations are compared to a more advanced approach based on a Riemannian metric. Very often the barycenter of the quaternions or matrices that represent the rotations are used as an estimate of the mean. These methods neglect that rotations belong ...... approximations to the Riemannian metric, and that the subsequent corrections are inherent in the least squares estimation.......In this paper two common approaches to averaging rotations are compared to a more advanced approach based on a Riemannian metric. Very often the barycenter of the quaternions or matrices that represent the rotations are used as an estimate of the mean. These methods neglect that rotations belong...

  16. A device for measuring electron beam characteristics

    Directory of Open Access Journals (Sweden)

    M. Andreev

    2017-01-01

    Full Text Available This paper presents a device intended for diagnostics of electron beams and the results obtained with this device. The device comprises a rotating double probe operating in conjunction with an automated probe signal collection and processing system. This provides for measuring and estimating the electron beam characteristics such as radius, current density, power density, convergence angle, and brightness.

  17. New irradiation devices at the FRN reactor

    International Nuclear Information System (INIS)

    Stark, W.

    1980-01-01

    In order to fulfill the experimental demands three additional devices were constructed and installed. The first is a vertical irradiation tube in air surrounded by a lead cylinder (in the irradiation position). The second device is a rabbit system ending within the graphite moderator of the thermal column. The third device is so called rotating disk assembly, built to replace the rotary specimen rack

  18. Tube coupling device

    Science.gov (United States)

    Myers, William N. (Inventor); Hein, Leopold A. (Inventor)

    1987-01-01

    A first annular ring of a tube coupling device has a keyed opening sized to fit around the nut region of a male coupling, and a second annular ring has a keyed opening sized to fit around the nut of a female coupling. Each ring has mating ratchet teeth and these rings are biased together, thereby engaging these teeth and preventing rotation of these rings. This in turn prevents the rotation of the male nut region with respect to the female nut. For tube-to-bulkhead locking, one facet of one ring is notched, and a pin is pressed into an opening in the bulkhead. This pin is sized to fit within one of the notches in the ring, thereby preventing rotation of this ring with respect to the bulkhead.

  19. Localized environment characterization device

    KAUST Repository

    Alzain, Hashim

    2016-07-21

    Various apparatuses and methods are provided for measuring the likely environmental impact of a particular geographic location on power generation properties of potential solar installations at the particular location. In an example embodiment of one such apparatus, a measurement device is provided. The measurement device includes a base portion comprising a base frame element disposed on a plurality of supporting legs, and a top panel comprising a series of connected members and one or more measurement modules whose planar dimensions are defined by the series of connected members. The top panel is connected to the base portion by a joint such that the top panel can rotate about the joint, and a panel support element is configured to fasten the top panel immovably at a desired degree of rotation in relation to the base portion.

  20. Detection device for inside of space

    International Nuclear Information System (INIS)

    Kojima, Kosuke.

    1996-01-01

    A securing plate is disposed to a support column which can be inserted to a space, and a rotational member on which equipments for detection are loaded is disposed rotatably in the longitudinal direction of the support column at a predetermined distance from the securing plate. Shape memory alloy wires that shrink when current is supplied are stretched circumferentially each at a predetermined distance obliquely relative to the support column between the rotational member and the securing plate. In addition, there are disposed a sensor for detecting the rotational angle of the rotational member, a calculator for determining the deviation of the angle of the rotation based on the difference of the detected rotational angle and a predetermined rotational angle as a reference, a deviation counter for obtaining deviation count signals based on the deviation of the rotational angle, a D-A convertor for converting the deviation count signals to analog signals, a pulse width modulation device for duty ratio control based on the analog signals and output pulse signals, and a power amplifier for amplifying the pulse signals and applying them to the shape memory wires. Since the device is reduced in the size and weight, handling is facilitated, and the rotational angle can be controlled accurately. The device can be used for detection of an end plate in a pressure vessel. (N.H.)

  1. Polarization splitter and polarization rotator designs based on transformation optics.

    Science.gov (United States)

    Kwon, Do-Hoon; Werner, Douglas H

    2008-11-10

    The transformation optics technique is employed in this paper to design two optical devices - a two-dimensional polarization splitter and a three-dimensional polarization rotator for propagating beams. The polarization splitter translates the TM- and the TE-polarized components of an incident beam in opposite directions (i.e., shifted up or shifted down). The polarization rotator rotates the polarization state of an incoming beam by an arbitrary angle. Both optical devices are reflectionless at the entry and exit interfaces. Design details and full-wave simulation results are provided.

  2. Nuclear fuel pellet charging device

    International Nuclear Information System (INIS)

    Komuro, Kojiro.

    1990-01-01

    The present invention concerns a nuclear fuel pellet loading device, in which nuclear fuel pellets are successively charged from an open end of a fuel can while rotating the can. That is, a fuel can sealed at one end with an end plug and opened at the other end is rotated around its pipe axis as the center on a rotationally diriving table. During rotation of the fuel can, nuclear fuel pellets are successively charged by means of a feed rod of a feeding device to the inside of the fuel can. The fuel can is rotated while being supported horizontally and the fuel pellets are charged from the open end thereof. Alternatively, the fuel can is rotated while being supported obliquely and the fuel pellets are charged gravitationally into the fuel can. In this way, the damages to the barrier of the fuel can can be reduce. Further, since the fuel pellets can be charged gravitationally by rotating the fuel can while being supported obliquely, the damages to the barrier can be reduced remarkably. (I.S.)

  3. Asteroid rotation rates

    International Nuclear Information System (INIS)

    Binzel, R.P.; Farinella, P.

    1989-01-01

    Within the last decade the data base of asteroid rotation parameters (rotation rates and lightcurve amplitudes) has become sufficiently large to identify some definite rends and properties which can help us to interpret asteroid collisional evolution. Many significant correlations are found between rotation parameters and diameter, with distinct changes occurring near 125 km. The size range, which is also the diameter above which self-gravity may become important, perhaps represents a division between surviving primordial asteroids and collisional fragments. A Maxwellian is able to fit the observed rotation rate distributions of asteroids with D>125 km, implying that their rotation rates may be determined by collisional evolution. Asteroids with D<125 km show an excess of slow rotators and their non-Maxwellian distributions suggests that their rotation rates are more strongly influenced by other processes, such as the distribution resulting from their formation in catastrophic disruption events. Other correlations observed in the data set include different mean rotation rates for C, S and M type asteroids implying that their surface spectra are indicative of bulk properties

  4. The spatial rotator

    DEFF Research Database (Denmark)

    Rasmusson, Allan; Hahn, Ute; Larsen, Jytte Overgaard

    2013-01-01

    This paper presents a new local volume estimator, the spatial rotator, which is based on measurements on a virtual 3D probe, using computer assisted microscopy. The basic design of the probe builds upon the rotator principle which requires only a few manual intersection markings, thus making...

  5. Superconducting rotating machines

    International Nuclear Information System (INIS)

    Smith, J.L. Jr.; Kirtley, J.L. Jr.; Thullen, P.

    1975-01-01

    The opportunities and limitations of the applications of superconductors in rotating electric machines are given. The relevant properties of superconductors and the fundamental requirements for rotating electric machines are discussed. The current state-of-the-art of superconducting machines is reviewed. Key problems, future developments and the long range potential of superconducting machines are assessed

  6. Fundamental Relativistic Rotator

    International Nuclear Information System (INIS)

    Staruszkiewicz, A.

    2008-01-01

    Professor Jan Weyssenhoff was Myron Mathisson's sponsor and collaborator. He introduced a class of objects known in Cracow as '' kreciolki Weyssenhoffa '', '' Weyssenhoff's rotating little beasts ''. The Author describes a particularly simple object from this class. The relativistic rotator described in the paper is such that its both Casimir invariants are parameters rather than constants of motion. (author)

  7. SMAP Faraday Rotation

    Science.gov (United States)

    Le Vine, David

    2016-01-01

    Faraday rotation is a change in the polarization as signal propagates through the ionosphere. At L-band it is necessary to correct for this change and measurements are made on the spacecraft of the rotation angle. These figures show that there is good agreement between the SMAP measurements (blue) and predictions based on models (red).

  8. Units of rotational information

    Science.gov (United States)

    Yang, Yuxiang; Chiribella, Giulio; Hu, Qinheping

    2017-12-01

    Entanglement in angular momentum degrees of freedom is a precious resource for quantum metrology and control. Here we study the conversions of this resource, focusing on Bell pairs of spin-J particles, where one particle is used to probe unknown rotations and the other particle is used as reference. When a large number of pairs are given, we show that every rotated spin-J Bell state can be reversibly converted into an equivalent number of rotated spin one-half Bell states, at a rate determined by the quantum Fisher information. This result provides the foundation for the definition of an elementary unit of information about rotations in space, which we call the Cartesian refbit. In the finite copy scenario, we design machines that approximately break down Bell states of higher spins into Cartesian refbits, as well as machines that approximately implement the inverse process. In addition, we establish a quantitative link between the conversion of Bell states and the simulation of unitary gates, showing that the fidelity of probabilistic state conversion provides upper and lower bounds on the fidelity of deterministic gate simulation. The result holds not only for rotation gates, but also to all sets of gates that form finite-dimensional representations of compact groups. For rotation gates, we show how rotations on a system of given spin can simulate rotations on a system of different spin.

  9. The rotating universe

    International Nuclear Information System (INIS)

    Ruben, G.; Treder, H.J.

    1987-01-01

    For a long time the question whether the universe rotates or not is discussed. Aspects of Huygens, Newton, Mach and other important historical scientists in this field are reported. The investigations of the mathematician Kurt Groedel in order to prove the rotation of the universe are illustrated. Kurt Groedel has shown that Einstein's gravitational equations of general relativity theory and the cosmological postulate of global homogeneity of cosmic matter (that is the Copernical principle) are not contradictionary to a rotating universe. Abberation measurements, position determination by means of radiointerferometry and methods for the determination of the rotation of the universe from the isotropy of the background radiation are presented. From these experiments it can be concluded that the universe seems not to rotate as already Einstein expected

  10. Rotating stars in relativity.

    Science.gov (United States)

    Paschalidis, Vasileios; Stergioulas, Nikolaos

    2017-01-01

    Rotating relativistic stars have been studied extensively in recent years, both theoretically and observationally, because of the information they might yield about the equation of state of matter at extremely high densities and because they are considered to be promising sources of gravitational waves. The latest theoretical understanding of rotating stars in relativity is reviewed in this updated article. The sections on equilibrium properties and on nonaxisymmetric oscillations and instabilities in f -modes and r -modes have been updated. Several new sections have been added on equilibria in modified theories of gravity, approximate universal relationships, the one-arm spiral instability, on analytic solutions for the exterior spacetime, rotating stars in LMXBs, rotating strange stars, and on rotating stars in numerical relativity including both hydrodynamic and magnetohydrodynamic studies of these objects.

  11. Technical guides/ methodological guide in case of tripping of a radioactivity detection gantry; Guides techniques / guide methodologique en cas de declenchement d'un portique de detection de radioactivite

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2004-07-01

    The most of operators of reprocessing plants or waste disposal plants have equipment to detect radiation as gantry. To answer to a clarification need procedures can be followed during the tripping of a gantry. Procedures have been made by cards, they follow the regulation in force in the area of radiation protection and environment protection. With these cards there are an annexe giving the terms noticed in the cards, a glossary on radioactivity as well a list of organisms approved by the government to proceed to control for the workers protection against the hazards of ionizing radiations. (N.C.)

  12. On Job Rotation

    OpenAIRE

    Metin M. Cosgel; Thomas J. Miceli

    1998-01-01

    A fundamental principle of economics with which Adam Smith begins The Wealth of Nations is the division of labor. Some firms, however, have been pursuing a practice called job rotation, which assigns each worker not to a single and specific task but to a set of several tasks among which he or she rotates with some frequency. We examine the practice of job rotation as a serious alternative to specialization, with three objectives. The first is to consider current and historical examples of job...

  13. Rotator cuff - self-care

    Science.gov (United States)

    ... this page: //medlineplus.gov/ency/patientinstructions/000358.htm Rotator cuff - self-care To use the sharing features on ... and shoulder exercises may help ease your symptoms. Rotator Cuff Problems Common rotator cuff problems include: Tendinitis , which ...

  14. Wiggler as spin rotators for RHIC

    International Nuclear Information System (INIS)

    Luccio, A.; Conte, M.

    1993-01-01

    The spin of a polarized particle in a circular accelerator can be rotated with an arrangement of dipoles with field mutually perpendicular and perpendicular to the orbit. To achieve spin rotation, a given field integral value is required. The device must be designed in a way that the particle orbit is distorted as little as possible. It is shown that wigglers with many periods are suitable to achieve spin rotation with minimum orbit distortions. Wigglers are also more compact than more established structures and will use less electric power. Additional advantages include their use for non distructive beam diagnostics. Results are given for the Relativistic Heavy Ion Collider (RHIC) in the polarized proton mode

  15. Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors.

    Science.gov (United States)

    Berbeco, Ross I; Jiang, Steve B; Sharp, Gregory C; Chen, George T; Mostafavi, Hassan; Shirato, Hiroki

    2004-01-21

    The design of an integrated radiotherapy imaging system (IRIS), consisting of gantry mounted diagnostic (kV) x-ray tubes and fast read-out flat-panel amorphous-silicon detectors, has been studied. The system is meant to be capable of three main functions: radiographs for three-dimensional (3D) patient set-up, cone-beam CT and real-time tumour/marker tracking. The goal of the current study is to determine whether one source/panel pair is sufficient for real-time tumour/marker tracking and, if two are needed, the optimal position of each relative to other components and the isocentre. A single gantry-mounted source/imager pair is certainly capable of the first two of the three functions listed above and may also be useful for the third, if combined with prior knowledge of the target's trajectory. This would be necessary because only motion in two dimensions is visible with a single imager/source system. However, with previously collected information about the trajectory, the third coordinate may be derived from the other two with sufficient accuracy to facilitate tracking. This deduction of the third coordinate can only be made if the 3D tumour/marker trajectory is consistent from fraction to fraction. The feasibility of tumour tracking with one source/imager pair has been theoretically examined here using measured lung marker trajectory data for seven patients from multiple treatment fractions. The patients' selection criteria include minimum mean amplitudes of the tumour motions greater than 1 cm peak-to-peak. The marker trajectory for each patient was modelled using the first fraction data. Then for the rest of the data, marker positions were derived from the imager projections at various gantry angles and compared with the measured tumour positions. Our results show that, due to the three dimensionality and irregular trajectory characteristics of tumour motion, on a fraction-to-fraction basis, a 'monoscopic' system (single source/imager) is inadequate for

  16. A discrete time-varying internal model-based approach for high precision tracking of a multi-axis servo gantry.

    Science.gov (United States)

    Zhang, Zhen; Yan, Peng; Jiang, Huan; Ye, Peiqing

    2014-09-01

    In this paper, we consider the discrete time-varying internal model-based control design for high precision tracking of complicated reference trajectories generated by time-varying systems. Based on a novel parallel time-varying internal model structure, asymptotic tracking conditions for the design of internal model units are developed, and a low order robust time-varying stabilizer is further synthesized. In a discrete time setting, the high precision tracking control architecture is deployed on a Voice Coil Motor (VCM) actuated servo gantry system, where numerical simulations and real time experimental results are provided, achieving the tracking errors around 3.5‰ for frequency-varying signals. Copyright © 2014 ISA. Published by Elsevier Ltd. All rights reserved.

  17. Rotating universe models

    International Nuclear Information System (INIS)

    Tozini, A.V.

    1984-01-01

    A review is made of some properties of the rotating Universe models. Godel's model is identified as a generalized filted model. Some properties of new solutions of the Einstein's equations, which are rotating non-stationary Universe models, are presented and analyzed. These models have the Godel's model as a particular case. Non-stationary cosmological models are found which are a generalization of the Godel's metrics in an analogous way in which Friedmann is to the Einstein's model. (L.C.) [pt

  18. Rotation Invariance Neural Network

    OpenAIRE

    Li, Shiyuan

    2017-01-01

    Rotation invariance and translation invariance have great values in image recognition tasks. In this paper, we bring a new architecture in convolutional neural network (CNN) named cyclic convolutional layer to achieve rotation invariance in 2-D symbol recognition. We can also get the position and orientation of the 2-D symbol by the network to achieve detection purpose for multiple non-overlap target. Last but not least, this architecture can achieve one-shot learning in some cases using thos...

  19. SU-E-T-521: Feasibility Study of a Rotational Step-And-Shoot IMRT Treatment Planning Approach

    International Nuclear Information System (INIS)

    Zhu, X; Chang, S; Cullip, T; Yuan, L; Zhang, X; Lian, J; Tang, X; Tracton, G; Dooley, J

    2014-01-01

    Purpose: Rotational step-and-shot IMRT (r-IMRT) could improve delivery efficiency with good dose conformity, especially if it can leverage the burst mode of the accelerator where radiation is turned on/off momentarily while the gantry rotates continuously. The challenge for the r-IMRT planning is to minimize the number of beams to achieve a fast and smooth rotational delivery. Methods: Treatment plans for r-IMRT were created using an in-house treatment planning system. To generate the plan using a very few beams, gantry angle was optimized by weighting the beam monitoring unit (MU), and beam shape optimization was a combination of column search with k-means clustering. A prostate case and a head and neck case were planned using r-IMRT. The dosimetry is compared to s-IMRT planned with Varian Eclipse treatment planning system. Results: With the same PTV dose coverage D95=100%, the r-IMRT plans shows comparable sparing as the s-IMRT plans in the prostate for the rectum D10cc and the bladder Dmean, and in the head and neck for the spinal cord Dmax, the brain stem Dmax, the left/right parotid Dmean, the larynx Dmean, and the mandible Dmean. Both plans meet the established institutional clinical dosimetric criteria. The r-IMRT plan uses 19 beam/405 MU for the prostate, and 68 beam/880 MU for the head and neck, while the s-IMRT uses 7 beam/724 MU and 9 beam/1812 MU, respectively. Compared to the corresponding s-IMRT, r-IMRT has a reduction of MUs of 44% for the prostate case and 41% for the head and neck case. Conclusions: We have demonstrated the feasibility of a rotational step and shoot IMRT treatment planning approach that significantly shortens the conventional IMRT treatment beam-on time without degrading the dose comformity

  20. SU-E-T-521: Feasibility Study of a Rotational Step-And-Shoot IMRT Treatment Planning Approach

    Energy Technology Data Exchange (ETDEWEB)

    Zhu, X [Univ. of North Carolina at Chapel Hill, Chapel Hill, NC (United States); Chang, S [UNC School of Medicine, Chapel Hill, NC (United States); Cullip, T [UNC Hospitals, Chapel Hill, NC (United States); Yuan, L; Zhang, X [Duke University, Durham, NC (United States); Lian, J; Tang, X [UniversityNorth Carolina, Chapel Hill, NC (United States); Tracton, G; Dooley, J [University of North Carolina, Chapel Hill, NC (United States)

    2014-06-01

    Purpose: Rotational step-and-shot IMRT (r-IMRT) could improve delivery efficiency with good dose conformity, especially if it can leverage the burst mode of the accelerator where radiation is turned on/off momentarily while the gantry rotates continuously. The challenge for the r-IMRT planning is to minimize the number of beams to achieve a fast and smooth rotational delivery. Methods: Treatment plans for r-IMRT were created using an in-house treatment planning system. To generate the plan using a very few beams, gantry angle was optimized by weighting the beam monitoring unit (MU), and beam shape optimization was a combination of column search with k-means clustering. A prostate case and a head and neck case were planned using r-IMRT. The dosimetry is compared to s-IMRT planned with Varian Eclipse treatment planning system. Results: With the same PTV dose coverage D95=100%, the r-IMRT plans shows comparable sparing as the s-IMRT plans in the prostate for the rectum D10cc and the bladder Dmean, and in the head and neck for the spinal cord Dmax, the brain stem Dmax, the left/right parotid Dmean, the larynx Dmean, and the mandible Dmean. Both plans meet the established institutional clinical dosimetric criteria. The r-IMRT plan uses 19 beam/405 MU for the prostate, and 68 beam/880 MU for the head and neck, while the s-IMRT uses 7 beam/724 MU and 9 beam/1812 MU, respectively. Compared to the corresponding s-IMRT, r-IMRT has a reduction of MUs of 44% for the prostate case and 41% for the head and neck case. Conclusions: We have demonstrated the feasibility of a rotational step and shoot IMRT treatment planning approach that significantly shortens the conventional IMRT treatment beam-on time without degrading the dose comformity.

  1. Sweeping-window arc therapy: an implementation of rotational IMRT with automatic beam-weight calculation

    International Nuclear Information System (INIS)

    Cameron, C

    2005-01-01

    Sweeping-window arc therapy (SWAT) is a variation of intensity-modulated radiation therapy (IMRT) with direct aperture optimization (DAO) that is initialized with a leaf sequence of sweeping windows that move back and forth periodically across the target as the gantry rotates. This initial sequence induces modulation in the dose and is assumed to be near enough to a minimum to allow successful optimization, done with simulated annealing, without requiring excessive leaf speeds. Optimal beam weights are calculated analytically, with easy extension to allow for variable beam weights. In this paper SWAT is tested on a phantom model and clinical prostate case. For the phantom, constant and variable beam weights are used. Although further work (in particular, improving the dose model) is required, the results show SWAT to be a feasible approach to generating deliverable dynamic arc treatments that are optimized

  2. Sweeping-window arc therapy: an implementation of rotational IMRT with automatic beam-weight calculation

    Energy Technology Data Exchange (ETDEWEB)

    Cameron, C [Division of Radiation Physics, Department of Radiation Oncology, Stanford Cancer Center, 875 Blake Wilbur Drive, Rm G-233, Stanford, CA 94305-5847 (United States)

    2005-09-21

    Sweeping-window arc therapy (SWAT) is a variation of intensity-modulated radiation therapy (IMRT) with direct aperture optimization (DAO) that is initialized with a leaf sequence of sweeping windows that move back and forth periodically across the target as the gantry rotates. This initial sequence induces modulation in the dose and is assumed to be near enough to a minimum to allow successful optimization, done with simulated annealing, without requiring excessive leaf speeds. Optimal beam weights are calculated analytically, with easy extension to allow for variable beam weights. In this paper SWAT is tested on a phantom model and clinical prostate case. For the phantom, constant and variable beam weights are used. Although further work (in particular, improving the dose model) is required, the results show SWAT to be a feasible approach to generating deliverable dynamic arc treatments that are optimized.

  3. Radioactive source manipulator and stowage device

    International Nuclear Information System (INIS)

    Burton, C.

    1980-01-01

    A description is given of a radioactive source manipulator and stowage device comprising: a cylindrical body; a transversely disposed socket formed near one end of said cylindrical body for receiving a radioactive source; a cylindrical sleeve rotatably mounted on said cylindrical body; and an aperture formed in the wall of said sleeve whereby rotation of said sleeve to axially align said aperture with said socket will permit a radioactive source to be inserted into and removed from said socket and rotation of said sleeve to move said aperture out of alignment with said socket when the socket contains a radioactive source readies the device for manipulation and stowage

  4. Under conditions of large geometric miss, tumor control probability can be higher for static gantry intensity-modulated radiation therapy compared to volume-modulated arc therapy for prostate cancer

    International Nuclear Information System (INIS)

    Balderson, Michael; Brown, Derek; Johnson, Patricia; Kirkby, Charles

    2016-01-01

    The purpose of this work was to compare static gantry intensity-modulated radiation therapy (IMRT) with volume-modulated arc therapy (VMAT) in terms of tumor control probability (TCP) under scenarios involving large geometric misses, i.e., those beyond what are accounted for when margin expansion is determined. Using a planning approach typical for these treatments, a linear-quadratic–based model for TCP was used to compare mean TCP values for a population of patients who experiences a geometric miss (i.e., systematic and random shifts of the clinical target volume within the planning target dose distribution). A Monte Carlo approach was used to account for the different biological sensitivities of a population of patients. Interestingly, for errors consisting of coplanar systematic target volume offsets and three-dimensional random offsets, static gantry IMRT appears to offer an advantage over VMAT in that larger shift errors are tolerated for the same mean TCP. For example, under the conditions simulated, erroneous systematic shifts of 15 mm directly between or directly into static gantry IMRT fields result in mean TCP values between 96% and 98%, whereas the same errors on VMAT plans result in mean TCP values between 45% and 74%. Random geometric shifts of the target volume were characterized using normal distributions in each Cartesian dimension. When the standard deviations were doubled from those values assumed in the derivation of the treatment margins, our model showed a 7% drop in mean TCP for the static gantry IMRT plans but a 20% drop in TCP for the VMAT plans. Although adding a margin for error to a clinical target volume is perhaps the best approach to account for expected geometric misses, this work suggests that static gantry IMRT may offer a treatment that is more tolerant to geometric miss errors than VMAT.

  5. Under conditions of large geometric miss, tumor control probability can be higher for static gantry intensity-modulated radiation therapy compared to volume-modulated arc therapy for prostate cancer.

    Science.gov (United States)

    Balderson, Michael; Brown, Derek; Johnson, Patricia; Kirkby, Charles

    2016-01-01

    The purpose of this work was to compare static gantry intensity-modulated radiation therapy (IMRT) with volume-modulated arc therapy (VMAT) in terms of tumor control probability (TCP) under scenarios involving large geometric misses, i.e., those beyond what are accounted for when margin expansion is determined. Using a planning approach typical for these treatments, a linear-quadratic-based model for TCP was used to compare mean TCP values for a population of patients who experiences a geometric miss (i.e., systematic and random shifts of the clinical target volume within the planning target dose distribution). A Monte Carlo approach was used to account for the different biological sensitivities of a population of patients. Interestingly, for errors consisting of coplanar systematic target volume offsets and three-dimensional random offsets, static gantry IMRT appears to offer an advantage over VMAT in that larger shift errors are tolerated for the same mean TCP. For example, under the conditions simulated, erroneous systematic shifts of 15mm directly between or directly into static gantry IMRT fields result in mean TCP values between 96% and 98%, whereas the same errors on VMAT plans result in mean TCP values between 45% and 74%. Random geometric shifts of the target volume were characterized using normal distributions in each Cartesian dimension. When the standard deviations were doubled from those values assumed in the derivation of the treatment margins, our model showed a 7% drop in mean TCP for the static gantry IMRT plans but a 20% drop in TCP for the VMAT plans. Although adding a margin for error to a clinical target volume is perhaps the best approach to account for expected geometric misses, this work suggests that static gantry IMRT may offer a treatment that is more tolerant to geometric miss errors than VMAT. Copyright © 2016 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

  6. Rotating positron tomographs revisited

    International Nuclear Information System (INIS)

    Townsend, D.; Defrise, M.; Geissbuhler, A.

    1994-01-01

    We have compared the performance of a PET scanner comprising two rotating arrays of detectors with that of the more conventional stationary-ring design. The same total number of detectors was used in each, and neither scanner had septa. For brain imaging, we find that the noise-equivalent count rate is greater for the rotating arrays by a factor of two. Rotating arrays have a sensitivity profile that peaks in the centre of the field of view, both axially and transaxially. In the transaxial plane, this effect offsets to a certain extent the decrease in the number of photons detected towards the centre of the brain due to self-absorption. We have also compared the performance of a rotating scanner to that of a full-ring scanner with the same number of rings. We find that a full-ring scanner with an axial extent of 16.2 cm (24 rings) is a factor of 3.5 more sensitive than a rotating scanner with 40% of the detectors and the same axial extent. (Author)

  7. The optical rotator

    DEFF Research Database (Denmark)

    Tandrup, T; Gundersen, Hans Jørgen Gottlieb; Jensen, Eva B. Vedel

    1997-01-01

    further discuss the methods derived from this principle and present two new local volume estimators. The optical rotator benefits from information obtained in all three dimensions in thick sections but avoids over-/ underprojection problems at the extremes of the cell. Using computer-assisted microscopes......The optical rotator is an unbiased, local stereological principle for estimation of cell volume and cell surface area in thick, transparent slabs, The underlying principle was first described in 1993 by Kieu Jensen (T. Microsc. 170, 45-51) who also derived an estimator of length, In this study we...... the extra measurements demand minimal extra effort and make this estimator even more efficient when it comes to estimation of individual cell size than many of the previous local estimators, We demonstrate the principle of the optical rotator in an example (the cells in the dorsal root ganglion of the rat...

  8. Vibrations of rotating machinery

    CERN Document Server

    Matsushita, Osami; Kanki, Hiroshi; Kobayashi, Masao; Keogh, Patrick

    2017-01-01

    This book opens with an explanation of the vibrations of a single degree-of-freedom (dof) system for all beginners. Subsequently, vibration analysis of multi-dof systems is explained by modal analysis. Mode synthesis modeling is then introduced for system reduction, which aids understanding in a simplified manner of how complicated rotors behave. Rotor balancing techniques are offered for rigid and flexible rotors through several examples. Consideration of gyroscopic influences on the rotordynamics is then provided and vibration evaluation of a rotor-bearing system is emphasized in terms of forward and backward whirl rotor motions through eigenvalue (natural frequency and damping ratio) analysis. In addition to these rotordynamics concerning rotating shaft vibration measured in a stationary reference frame, blade vibrations are analyzed with Coriolis forces expressed in a rotating reference frame. Other phenomena that may be assessed in stationary and rotating reference frames include stability characteristic...

  9. The use of rotating electric arc for spherical particle production

    International Nuclear Information System (INIS)

    Bica, I.

    2000-01-01

    This work presents and experimental device designed to obtain spherical particles by mans of a rotating electric arc. A rotation frequency of the electric arc of 750 s''-1, a voltage of 50 V(dc) and a current of 100 A was used. The mass flow rate was 3 g.min''-1. Under these conditions particles of 15 to 20 μm in diameter were obtained. (Author) 8 refs

  10. Rotating quantum states

    International Nuclear Information System (INIS)

    Ambruş, Victor E.; Winstanley, Elizabeth

    2014-01-01

    We revisit the definition of rotating thermal states for scalar and fermion fields in unbounded Minkowski space–time. For scalar fields such states are ill-defined everywhere, but for fermion fields an appropriate definition of the vacuum gives thermal states regular inside the speed-of-light surface. For a massless fermion field, we derive analytic expressions for the thermal expectation values of the fermion current and stress–energy tensor. These expressions may provide qualitative insights into the behaviour of thermal rotating states on more complex space–time geometries

  11. Rotating bubble membrane radiator

    Science.gov (United States)

    Webb, Brent J.; Coomes, Edmund P.

    1988-12-06

    A heat radiator useful for expelling waste heat from a power generating system aboard a space vehicle is disclosed. Liquid to be cooled is passed to the interior of a rotating bubble membrane radiator, where it is sprayed into the interior of the bubble. Liquid impacting upon the interior surface of the bubble is cooled and the heat radiated from the outer surface of the membrane. Cooled liquid is collected by the action of centrifical force about the equator of the rotating membrane and returned to the power system. Details regarding a complete space power system employing the radiator are given.

  12. Bio-inspired optical rotation sensor

    Science.gov (United States)

    O'Carroll, David C.; Shoemaker, Patrick A.; Brinkworth, Russell S. A.

    2007-01-01

    Traditional approaches to calculating self-motion from visual information in artificial devices have generally relied on object identification and/or correlation of image sections between successive frames. Such calculations are computationally expensive and real-time digital implementation requires powerful processors. In contrast flies arrive at essentially the same outcome, the estimation of self-motion, in a much smaller package using vastly less power. Despite the potential advantages and a few notable successes, few neuromorphic analog VLSI devices based on biological vision have been employed in practical applications to date. This paper describes a hardware implementation in aVLSI of our recently developed adaptive model for motion detection. The chip integrates motion over a linear array of local motion processors to give a single voltage output. Although the device lacks on-chip photodetectors, it includes bias circuits to use currents from external photodiodes, and we have integrated it with a ring-array of 40 photodiodes to form a visual rotation sensor. The ring configuration reduces pattern noise and combined with the pixel-wise adaptive characteristic of the underlying circuitry, permits a robust output that is proportional to image rotational velocity over a large range of speeds, and is largely independent of either mean luminance or the spatial structure of the image viewed. In principle, such devices could be used as an element of a velocity-based servo to replace or augment inertial guidance systems in applications such as mUAVs.

  13. Photovoltaic device

    Energy Technology Data Exchange (ETDEWEB)

    Reese, Jason A; Keenihan, James R; Gaston, Ryan S; Kauffmann, Keith L; Langmaid, Joseph A; Lopez, Leonardo; Maak, Kevin D; Mills, Michael E; Ramesh, Narayan; Teli, Samar R

    2017-03-21

    The present invention is premised upon an improved photovoltaic device ("PV device"), more particularly to an improved photovoltaic device with a multilayered photovoltaic cell assembly and a body portion joined at an interface region and including an intermediate layer, at least one interconnecting structural member, relieving feature, unique component geometry, or any combination thereof.

  14. Photovoltaic device

    Science.gov (United States)

    Reese, Jason A.; Keenihan, James R.; Gaston, Ryan S.; Kauffmann, Keith L.; Langmaid, Joseph A.; Lopez, Leonardo C.; Maak, Kevin D.; Mills, Michael E.; Ramesh, Narayan; Teli, Samar R.

    2015-06-02

    The present invention is premised upon an improved photovoltaic device ("PV device"), more particularly to an improved photovoltaic device with a multilayered photovoltaic cell assembly and a body portion joined at an interface region and including an intermediate layer, at least one interconnecting structural member, relieving feature, unique component geometry, or any combination thereof.

  15. Photovoltaic device

    Science.gov (United States)

    Reese, Jason A.; Keenihan, James R.; Gaston, Ryan S.; Kauffmann, Keith L.; Langmaid, Joseph A.; Lopez, Leonardo C.; Maak, Kevin D.; Mills, Michael E.; Ramesh, Narayan; Teli, Samar R.

    2015-09-01

    The present invention is premised upon an improved photovoltaic device ("PV device"), more particularly to an improved photovoltaic device (10) with a multilayered photovoltaic cell assembly (100) and a body portion (200) joined at an interface region (410) and including an intermediate layer (500), at least one interconnecting structural member (1500), relieving feature (2500), unique component geometry, or any combination thereof.

  16. Rotational manipulation of single cells and organisms using acoustic waves.

    Science.gov (United States)

    Ahmed, Daniel; Ozcelik, Adem; Bojanala, Nagagireesh; Nama, Nitesh; Upadhyay, Awani; Chen, Yuchao; Hanna-Rose, Wendy; Huang, Tony Jun

    2016-03-23

    The precise rotational manipulation of single cells or organisms is invaluable to many applications in biology, chemistry, physics and medicine. In this article, we describe an acoustic-based, on-chip manipulation method that can rotate single microparticles, cells and organisms. To achieve this, we trapped microbubbles within predefined sidewall microcavities inside a microchannel. In an acoustic field, trapped microbubbles were driven into oscillatory motion generating steady microvortices which were utilized to precisely rotate colloids, cells and entire organisms (that is, C. elegans). We have tested the capabilities of our method by analysing reproductive system pathologies and nervous system morphology in C. elegans. Using our device, we revealed the underlying abnormal cell fusion causing defective vulval morphology in mutant worms. Our acoustofluidic rotational manipulation (ARM) technique is an easy-to-use, compact, and biocompatible method, permitting rotation regardless of optical, magnetic or electrical properties of the sample under investigation.

  17. SU-E-J-138: An IGRT QA Device for Measuring with Tenths-Millimeter Accuracy KV and MV Isocenter Congruence, Couch Travel and Laser Alignment of Accelerators Used for SRS and SBRT

    Energy Technology Data Exchange (ETDEWEB)

    Brezovich, I; Popple, R; Duan, J; Huang, M; Benhabib, S; Shen, S; Cardan, R; Wu, X [University of Alabama Birmingham, Birmingham, AL (United States)

    2014-06-01

    Purpose: To develop a practical device having sufficient accuracy for daily QA tests of accelerators used for SRS and SBRT. Methods: The UAB (Universal Alignment Ball) consists of a 6.35 mm (1/4 inch) diameter tungsten sphere located concentrically within a 25.4 mm (1 inch) diameter acrylic plastic (PMMA) sphere. The spheres are embedded in polystyrene foam, which, in turn, is surrounded by a cylindrical PMMA shell. The UAB is placed on the couch and aligned with wall lasers according to marks that have known positions in relation to the center of the spheres. Using planar and cone beam images the couch is shifted till the surface of the PMMA sphere matches Eclipse-generated circular contours. Anterior and lateral MV images taken with small MLC openings allow measurement of distance between kV and MV isocenter, laser and MLC alignment. Measurements were taken over a one-month period. Results: Artifacts from the tungsten sphere were confined within the PMMA sphere and did not affect cone beam localization of the sphere boundary, allowing 0.1 mm precise alignment with a computer-generated circle centered at kV isocenter. In tests extending over a one-month period, the distance between kV and MV isocenters along the vertical, longitudinal and lateral directions was 0.125 +/−0.06, 0.19 +/−0.08, and 0.02 +/−0.08 mm, respectively. Laser misalignment along these directions was 0.34 +/- 0.15, 0.74 +/−0.29, and 0.49 +/−0.22 mm. Automated couch shifts moved the spheres to within 0.1 mm of the selected position. The center of a 1cmx1cm MLC-defined field remained within +/−0.2 mm of the tungsten sphere center as the gantry was rotated. Conclusion: The UAB is practical for daily end-to-end QA tests of accelerator alignment. It provides tenths-mm accuracy for measuring agreement of kV and MV isocenters, couch motions, gantry flex and laser alignment.

  18. Continuous Arc Rotation of the Couch Therapy for the Delivery of Accelerated Partial Breast Irradiation: A Treatment Planning Analysis

    International Nuclear Information System (INIS)

    Shaitelman, Simona F.; Kim, Leonard H.; Yan Di; Martinez, Alvaro A.; Vicini, Frank A.; Grills, Inga S.

    2011-01-01

    Purpose: We present a novel form of arc therapy: continuous arc rotation of the couch (C-ARC) and compare its dosimetry with three-dimensional conformal radiotherapy (3D-CRT), intensity-modulated radiotherapy (IMRT), and volumetric-modulated arc therapy (VMAT) for accelerated partial breast irradiation (APBI). C-ARC, like VMAT, uses a modulated beam aperture and dose rate, but with the couch, not the gantry, rotating. Methods and Materials: Twelve patients previously treated with APBI using 3D-CRT were replanned with (1) C-ARC, (2) IMRT, and (3) VMAT. C-ARC plans were designed with one medial and one lateral arc through which the couch rotated while the gantry was held stationary at a tangent angle. Target dose coverage was normalized to the 3D-CRT plan. Comparative endpoints were dose to normal breast tissue, lungs, and heart and monitor units prescribed. Results: Compared with 3D-CRT, C-ARC, IMRT, and VMAT all significantly reduced the ipsilateral breast V50% by the same amount (mean, 7.8%). Only C-ARC and IMRT plans significantly reduced the contralateral breast maximum dose, the ipsilateral lung V5Gy, and the heart V5%. C-ARC used on average 40%, 30%, and 10% fewer monitor units compared with 3D-CRT, IMRT, and VMAT, respectively. Conclusions: C-ARC provides improved dosimetry and treatment efficiency, which should reduce the risks of toxicity and secondary malignancy. Its tangent geometry avoids irradiation of critical structures that is unavoidable using the en face geometry of VMAT.

  19. Rotations and angular momentum

    International Nuclear Information System (INIS)

    Nyborg, P.; Froyland, J.

    1979-01-01

    This paper is devoted to the analysis of rotational invariance and the properties of angular momentum in quantum mechanics. In particular, the problem of addition of angular momenta is treated in detail, and tables of Clebsch-Gordan coefficients are included

  20. Negative Rotation Cinch Strap.

    Science.gov (United States)

    This project discloses an improved unitary parachute torso harness, having a single fastening means, wherein an auxillary tightening strap is...attached to the groin straps of said harness. Said auxillary straps are used to prevent torso rotation or harness slippage and to prevent harness elongation

  1. A rotating string

    International Nuclear Information System (INIS)

    Jensen, B.

    1993-06-01

    The author presents a global solution of Einstein's equations which represents a rotating cosmic string with a finite coreradius. The importance of pressure for the generation of closed timelike curves outside the coreregion of such strings is clearly displayed in this model due to the simplicity of the source. 10 refs

  2. Rotator Cuff Injuries.

    Science.gov (United States)

    Connors, G. Patrick

    Many baseball players suffer from shoulder injuries related to the rotator cuff muscles. These injuries may be classified as muscular strain, tendonitis or tenosynovitis, and impingement syndrome. Treatment varies from simple rest to surgery, so it is important to be seen by a physician as soon as possible. In order to prevent these injuries, the…

  3. Rotational dynamics with Tracker

    International Nuclear Information System (INIS)

    Eadkhong, T; Danworaphong, S; Rajsadorn, R; Jannual, P

    2012-01-01

    We propose the use of Tracker, freeware for video analysis, to analyse the moment of inertia (I) of a cylindrical plate. Three experiments are performed to validate the proposed method. The first experiment is dedicated to find the linear coefficient of rotational friction (b) for our system. By omitting the effect of such friction, we derive I for a cylindrical plate rotated around its central axis from the other two experiments based on the relation between torque and angular acceleration of rotational motion and conservation of energy. Movies of the rotating plate and hung masses are recorded. As a result, we have the deviation of I from its theoretical value of 0.4% and 3.3%, respectively. Our setup is completely constructed from locally available inexpensive materials and the experimental results indicate that the system is highly reliable. This work should pave the way for those who prefer to build a similar setup from scratch at relatively low cost compared to commercial units. (paper)

  4. Can planetary nebulae rotate

    International Nuclear Information System (INIS)

    Grinin, V.P.

    1982-01-01

    It is shown that the inclination of spectral lines observed in a number of planetary nebulae when the spectrograph slit is placed along the major axis, which is presently ascribed to nonuniform expansion of the shells, actually may be due to rotation of the nebulae about their minor axes, as Campbell and Moore have suggested in their reports. It is assumed that the rotation of the central star (or, if the core is a binary system, circular motions of gas along quasi-Keplerian orbits) serves as the source of the original rotation of a protoplanetary nebula. The mechanism providing for strengthening of the original rotation in the process of expansion of the shell is the tangential pressure of L/sub α/ radiation due to the anisotropic properties of the medium and radiation field. The dynamic effect produced by them is evidently greatest in the epoch when the optical depth of the nebula in the L/sub c/ continuum becomes on the order of unity in the course of its expansion

  5. Device for contamination monitoring against radiation contamination of people

    International Nuclear Information System (INIS)

    Rische, U.W.; Gerlach, R.

    1986-01-01

    The monitor has detector devices at an angle to each other made as a rigid component which can be rotated around a vertical axis in the angle between the joined detector devices. A reset drive which can be tensioned is provided at the axis of rotation. If it is in its rest position, a platform is situated as floor plate with a foot detector between the vertical detector devices. (orig./HP) [de

  6. Wave induced supersonic rotation in mirrors

    Science.gov (United States)

    Fetterman, Abraham

    2010-11-01

    Wave-particle interactions in ExB supersonically rotating plasmas feature an unusual effect: particles are diffused by waves in both potential energy and kinetic energy [1]. This wave-particle interaction generalizes the alpha channeling effect, in which radio frequency waves are used to remove alpha particles collisionlessly at low energy. In rotating plasmas, the alpha particles may be removed at low energy through the loss cone, and the energy lost may be transferred to the radial electric field. This eliminates the need for electrodes in the mirror throat, which have presented serious technical issues in past rotating plasma devices. A particularly simple way to achieve this effect is to use a high azimuthal mode number perturbation on the magnetic field [2]. In the rotating frame, this perturbation is seen as a wave near the alpha particle cyclotron harmonic, and can break the azimuthal symmetry and magnetic moment conservation without changing the particle's total energy. The particle may exit if it reduces its kinetic energy and becomes more trapped if it gains kinetic energy, leading to a steady state current that maintains the field. Simulations of single particles in rotating mirrors show that a stationary wave can extract enough energy from alpha particles for a reactor to be self-sustaining. Rotation can also be sustained by waves in plasmas without a kinetic energy source. This type of wave has been considered for plasma centrifuges used for isotope separation [3]. [4pt] [1] A. J. Fetterman and N. J. Fisch, Phys Rev Lett 101, 205003 (2008). [0pt] [2] A. J. Fetterman and N. J. Fisch, Phys. Plasmas 17, 042112 (2010). [0pt] [3] A. J. Fetterman and N. J. Fisch, Plasma Sources Sci. Tech. 18, 045003 (2009).

  7. SU-F-J-191: Dosimetric Evaluation of a Left Chestwall Patient Treated with a Compact Proton Pencil Beam Gantry Utilizing Daily Setup CBCT

    Energy Technology Data Exchange (ETDEWEB)

    Maynard, M; Chen, K; Rosen, L; Wu, H [Willis-Knighton Medical Center, Shreveport, LA (United States)

    2016-06-15

    Purpose: To evaluate the robustness of the gradient technique for treating a multi-isocenter left chest wall patient with a compact proton pencil beam gantry. Both CBCT and stereoscopic imaging are used to facilitate daily treatment setup. Methods: To treat the elongated chest wall planning target volume (PTV) with the compact PBS system, a 28 fraction (5040 CcGE) treatment plan was created using two fields with gradient matching technique. Daily table shifts between treatment field isocenters were obtained from the record and verify system for each treatment fraction. Copies of the initial treatment plan were made for each fraction and the field isocenter coordinates for each plan copy were adjusted to reflect daily table shifts. Doses were re-calculated for each fraction, summed, and compared against the initial plan. Results: The table shifts (average and range) were 2.2 (−5.1–+3.9), 3.0 (−6.0–+4.0) and 3.0 (−10.1–+1.9) millimeters in the anterior-posterior, superior-inferior and right-left directions, respectively. Dose difference to the PTV, heart and ipsilateral lung were evaluated. The percentage of the PTV receiving the prescription dose decreased from 94.6% to 89.1%. The D95 of the PTV increased from 99.6% to 99.9%. The maximum dose in PTV increased from 106.6% to 109.2% and V105 increased from 1.0% to 16.5%. The V20 of the ipsilateral lung increased from 18.5% to 21.0%. The mean heart dose difference was negligible. Conclusion: Observed dose differences to lung and heart tissues due to daily setup variations remained acceptably low while maintaining sufficient dose coverage to the PTV. This initial case study demonstrates the robustness of the gradient technique to treat a large target, multi-isocenter plan with a compact proton pencil beam gantry equipped with CBCT and stereoscopic imaging modalities.

  8. Faraday rotation measurement method and apparatus

    Science.gov (United States)

    Brockman, M. H. (Inventor)

    1981-01-01

    A method and device for measuring Faraday rotation of a received RF signal is described. A simultaneous orthogonal polarization receiver compensates for a 3 db loss due to splitting of a received signal into left circular and right circular polarization channels. The compensation is achieved by RF and modulation arraying utilizing a specific receiver array which also detects and measures Faraday rotation in the presence or absence of spin stabilization effects on a linear polarization vector. Either up-link or down-link measurement of Faraday rotation is possible. Specifically, the Faraday measurement apparatus utilized in conjunction with the specific receiver array provides a means for comparing the phase of a reference signal in the receiver array to the phase of a tracking loop signal related to the incoming signal, and comparing the phase of the reference signal to the phase of the tracking signal shifted in phase by 90 degrees. The averaged and unaveraged signals, are compared, the phase changes between the two signals being related to Faraday rotation.

  9. Wave-particle interactions in rotating mirrorsa)

    Science.gov (United States)

    Fetterman, Abraham J.; Fisch, Nathaniel J.

    2011-05-01

    Wave-particle interactions in E ×B rotating plasmas feature an unusual effect: particles are diffused by waves in both potential energy and kinetic energy. This wave-particle interaction generalizes the alpha channeling effect, in which radio frequency waves are used to remove alpha particles collisionlessly at low energy. In rotating plasmas, the alpha particles may be removed at low energy through the loss cone, and the energy lost may be transferred to the radial electric field. This eliminates the need for electrodes in the mirror throat, which have presented serious technical issues in past rotating plasma devices. A particularly simple way to achieve this effect is to use a high azimuthal mode number perturbation on the magnetic field. Rotation can also be sustained by waves in plasmas without a kinetic energy source. This type of wave has been considered for plasma centrifuges used for isotope separation. Energy may also be transferred from the electric field to particles or waves, which may be useful for ion heating and energy generation.

  10. Wave-particle Interactions In Rotating Mirrors

    Energy Technology Data Exchange (ETDEWEB)

    Abraham J. Fetterman and Nathaniel J. Fisch

    2011-01-11

    Wave-particle interactions in E×B rotating plasmas feature an unusual effect: particles are diffused by waves in both potential energy and kinetic energy. This wave-particle interaction generalizes the alpha channeling effect, in which radio frequency waves are used to remove alpha particles collisionlessly at low energy. In rotating plasmas, the alpha particles may be removed at low energy through the loss cone, and the energy lost may be transferred to the radial electric field. This eliminates the need for electrodes in the mirror throat, which have presented serious technical issues in past rotating plasma devices. A particularly simple way to achieve this effect is to use a high azimuthal mode number perturbation on the magnetic field. Rotation can also be sustained by waves in plasmas without a kinetic energy source. This type of wave has been considered for plasma centrifuges used for isotope separation. Energy may also be transferred from the electric field to particles or waves, which may be useful for ion heating and energy generation.

  11. Wave-particle interactions in rotating mirrors

    International Nuclear Information System (INIS)

    Fetterman, Abraham J.; Fisch, Nathaniel J.

    2011-01-01

    Wave-particle interactions in ExB rotating plasmas feature an unusual effect: particles are diffused by waves in both potential energy and kinetic energy. This wave-particle interaction generalizes the alpha channeling effect, in which radio frequency waves are used to remove alpha particles collisionlessly at low energy. In rotating plasmas, the alpha particles may be removed at low energy through the loss cone, and the energy lost may be transferred to the radial electric field. This eliminates the need for electrodes in the mirror throat, which have presented serious technical issues in past rotating plasma devices. A particularly simple way to achieve this effect is to use a high azimuthal mode number perturbation on the magnetic field. Rotation can also be sustained by waves in plasmas without a kinetic energy source. This type of wave has been considered for plasma centrifuges used for isotope separation. Energy may also be transferred from the electric field to particles or waves, which may be useful for ion heating and energy generation.

  12. Wave-particle Interactions In Rotating Mirrors

    International Nuclear Information System (INIS)

    Fetterman, Abraham J.; Fisch, Nathaniel J.

    2011-01-01

    Wave-particle interactions in E-B rotating plasmas feature an unusual effect: particles are diffused by waves in both potential energy and kinetic energy. This wave-particle interaction generalizes the alpha channeling effect, in which radio frequency waves are used to remove alpha particles collisionlessly at low energy. In rotating plasmas, the alpha particles may be removed at low energy through the loss cone, and the energy lost may be transferred to the radial electric field. This eliminates the need for electrodes in the mirror throat, which have presented serious technical issues in past rotating plasma devices. A particularly simple way to achieve this effect is to use a high azimuthal mode number perturbation on the magnetic field. Rotation can also be sustained by waves in plasmas without a kinetic energy source. This type of wave has been considered for plasma centrifuges used for isotope separation. Energy may also be transferred from the electric field to particles or waves, which may be useful for ion heating and energy generation.

  13. Handling device for nuclear fuel assemblies and assembly appropriate for such a device

    International Nuclear Information System (INIS)

    Cransac, J.P.; Jaquelin, R.; Renaux, C.

    1985-01-01

    The handling device comprises a guide tube of which axis is vertical, in which a grab moves, hanging from a chain, under the action of a back-geared motor. The grab being stopped in its rotation in the guide tube, an assembly can be gripped with a bayonet system while controlling the rotation of the grab - guide tube system a back-geared motor. The device can be hanged from the small or large rotating plug of a fast neutron reactor. It can be used in a handling flask [fr

  14. Wave-driven Rotation in Supersonically Rotating Mirrors

    Energy Technology Data Exchange (ETDEWEB)

    A. Fetterman and N.J. Fisch

    2010-02-15

    Supersonic rotation in mirrors may be produced by radio frequency waves. The waves produce coupled diffusion in ion kinetic and potential energy. A population inversion along the diffusion path then produces rotation. Waves may be designed to exploit a natural kinetic energy source or may provide the rotation energy on their own. Centrifugal traps for fusion and isotope separation may benefit from this wave-driven rotation.

  15. Wave-driven Rotation in Supersonically Rotating Mirrors

    International Nuclear Information System (INIS)

    Fetterman, A.; Fisch, N.J.

    2010-01-01

    Supersonic rotation in mirrors may be produced by radio frequency waves. The waves produce coupled diffusion in ion kinetic and potential energy. A population inversion along the diffusion path then produces rotation. Waves may be designed to exploit a natural kinetic energy source or may provide the rotation energy on their own. Centrifugal traps for fusion and isotope separation may benefit from this wave-driven rotation.

  16. Governor device

    International Nuclear Information System (INIS)

    Domagala, J.E.; Clements, D.J.; Prior, P.C.S.

    1985-01-01

    A governor for limiting the speed of a rotating member relative to another member, has one member in the form of a conducting disc, and the other incorporates a plurality of magnets disposed circumferentially in an appropriate polarity pattern so that relative rotation results in the generation of a braking torque. In a modification the members are movable axially to vary the braking torque developed. The gap is preferably reduced automatically to increase the braking torque when the relative speeds attain a predetermined high value. (author)

  17. Sealing device

    Science.gov (United States)

    Garcia-Crespo, Andres Jose

    2013-12-10

    A sealing device for sealing a gap between a dovetail of a bucket assembly and a rotor wheel is disclosed. The sealing device includes a cover plate configured to cover the gap and a retention member protruding from the cover plate and configured to engage the dovetail. The sealing device provides a seal against the gap when the bucket assemply is subjected to a centrifugal force.

  18. Microfluidic Device

    Science.gov (United States)

    Tai, Yu-Chong (Inventor); Zheng, Siyang (Inventor); Lin, Jeffrey Chun-Hui (Inventor); Kasdan, Harvey L. (Inventor)

    2017-01-01

    Described herein are particular embodiments relating to a microfluidic device that may be utilized for cell sensing, counting, and/or sorting. Particular aspects relate to a microfabricated device that is capable of differentiating single cell types from dense cell populations. One particular embodiment relates a device and methods of using the same for sensing, counting, and/or sorting leukocytes from whole, undiluted blood samples.

  19. Gauging device

    International Nuclear Information System (INIS)

    Qurnell, F.D.; Patterson, C.B.

    1979-01-01

    A gauge supporting device for measuring say a square tube comprises a pair of rods or guides in tension between a pair of end members, the end members being spaced apart by a compression member or members. The tensioned guides provide planes of reference for measuring devices moved therealong on a carriage. The device is especially useful for making on site dimensional measurements of components, such as irradiated and therefore radioactive components, that cannot readily be transported to an inspection laboratory. (UK)

  20. Radioactive substance removing device

    International Nuclear Information System (INIS)

    Takeuchi, Jun; Tayama, Ryuichi; Teruyama, Hidehiko; Hikichi, Takayoshi.

    1992-01-01

    If inert gases are jetted from a jetting device to liquid metals in a capturing vessel, the inert gases are impinged on the inner wall surface of the capturing vessel, to reduce the thickness of a boundary layer as a diffusion region of radioactive materials formed between the inner wall surface of the capturing vessel and the liquid metals. Further, a portion of the boundary layer is peeled off to increase the adsorption amount of radioactive materials by the capturing vessel. When the inert gases are jetted on the inner or outer circumference of the capturing vessel to rotate the capturing vessel, the flow of the liquid metals is formed along with the rotation, and the thickness of the boundary layer is reduced or the boundary layer is peeled off to increase the absorption amount of the radioactive materials. If gas bubbles are formed in the liquid metals by the inert gases, the liquid metals are stirred by the gas bubbles to reduce the thickness of the boundary layer or peel it off, thereby enabling to increase the adsorption amount of the radioactive materials. Since it is not necessary to pass through the rotational member to the wall surface of the vessel, safety and reliability can be improved. (N.H.)

  1. Plasma rotation study in Tore Supra radio frequency heated plasmas

    International Nuclear Information System (INIS)

    Chouli, Bilal

    2014-01-01

    Toroidal flows are found to improve the performance of the magnetic confinement devices with increase of the plasma stability and confinement. In ITER or future reactors, the torque from NBI should be less important than in present-day tokamaks. Consequently, it is of interest to study other intrinsic mechanisms that can give rise to plasma rotation in order to predict the rotation profile in experiments. Intriguing observations of plasmas rotation have been made in radio frequency (RF) heated plasmas with little or no external momentum injection. Toroidal rotation in both the direction of the plasma current (co-current) and in the opposite direction (counter-current) has been observed depending on the heating schemes and plasma performance. In Tore Supra, most observations in L-mode plasmas have been in the counter-current direction. However, in this thesis, we show that in lower hybrid current drive (LHCD), the core toroidal rotation increment is in co- or counter-current direction depending on the plasma current amplitude. At low plasma current the rotation change is in the co-current direction while at high plasma current, the change is in the counter-current direction. In both low and high plasma current cases, rotation increments are found to increase linearly with the injected LH power. Several mechanisms in competition which can induce co- or counter-current rotation in Tore Supra LHCD plasmas are investigated and typical order of magnitude are discussed in this thesis. (author) [fr

  2. Rotator Cuff Injuries - Multiple Languages

    Science.gov (United States)

    ... Are Here: Home → Multiple Languages → All Health Topics → Rotator Cuff Injuries URL of this page: https://medlineplus.gov/ ... V W XYZ List of All Topics All Rotator Cuff Injuries - Multiple Languages To use the sharing features ...

  3. Numerical simulation of the temperature effects on the performance of rotational supercavitating evaporator

    International Nuclear Information System (INIS)

    Zheng, Z Y; Cheng, J P; Li, F C; Zhang, M; Li, Q; Kulagin, V A

    2015-01-01

    With the application of supercavitation effect, a novel device named Rotational Supercavitating Evaporator (RSCE) has been designed for desalination. In order to study the effect of temperature on the performance of RSCE and then direct the experimental study on RSCE for the next step, numerical simulations are conducted on the supercavitating flows in RSCE under different temperatures and rotational speeds. The results show that the rotational speed, resistance moment and mechanical energy consumed by the rotational cavitator under the critical state with the largest supercavity, decrease with the increase of temperature. And the area and volume of the supercavity increase exponentially with the increase of temperature under the same rotational speed

  4. Rotating flux compressor for energy conversion

    International Nuclear Information System (INIS)

    Chowdhuri, P.; Linton, T.W.; Phillips, J.A.

    1983-01-01

    The rotating flux compressor (RFC) converts rotational kinetic energy into an electrical output pulse which would have higher energy than the electrical energy initially stored in the compressor. An RFC has been designed in which wedge-shaped rotor blades pass through the air gaps between successive turns of a solenoid, the stator. Magnetic flux is generated by pulsing the stator solenoids when the inductance is a maximum, i.e., when the flux fills the stator-solenoid volume. Connecting the solenoid across a load conserves the flux which is compressed within the small volume surrounding the stator periphery when the rotor blades cut into the free space between the stator plates, creating a minimum-inductance condition. The unique features of this design are: (1) no electrical connections (brushes) to the rotor; (2) no conventional windings; and (3) no maintenance. The device has been tested up to 5000 rpm of rotor speed

  5. Asteroid rotation. IV

    International Nuclear Information System (INIS)

    Harris, A.W.; Young, J.W.

    1983-01-01

    The results from the year 1979 of an ongoing program of asteroid photometry at Table Mountain Observatory are presented. The results for 53 asteroids are summarized in a table, showing the number, name, opposition date, taxonomic class, diameter, absolute magnitude, mean absolute magnitude at zero phase angle and values of the absolute magnitude and linear phase coefficient derived from it, the rotation period in hours, peak-to-peak amplitude of variation, difference between mean and maximum brightness, and reliability index. Another table presents data on aspect and comparison stars, including brightness and distance data. Reliable rotation periods are reported for 22 asteroids for which no previous values are known. For seven asteroids, periods are reported which are revisions of previously reported values

  6. Rotator cuff disease

    International Nuclear Information System (INIS)

    Ziatkin, M.B.; Iannotti, J.P.; Roberts, M.; Dalinka, M.K.; Esterhai, J.L.; Kressel, H.Y.; Lenkinski, R.E.

    1988-01-01

    A dual-surface-coil array in a Helmholtz configuration was used to evaluate th rotator cuff in ten normal volunteers and 44 patients. Studies were performed with a General Electric 1.5-T MR imager. Thirty-two patients underwent surgery, 25 of whom also underwent arthrography. In comparison with surgery, MR imaging was more sensitive than arthrography for rotator cuff tears (91% vs 71%). The specificity and accuracy of MR imaging were 88% and 91%. The accuracy increased with use of an MR grading system. MR findings correlated with surgical findings with regard to the size and site of tears. MR findings of cuff tears were studied with multivariate analysis. Correlation was also found between a clinical score, the MR grade, and the clinical outcome

  7. The Spatiale Rotator

    DEFF Research Database (Denmark)

    Rasmusson, Allan

    2009-01-01

    it is embedded and sectioned. This has the unfortunate side effect that all information about positioning within the object is lost for blocks and sections. For complex tissue, like the mammalian brain, this information is of utmost importance to ensure measurements are performed in the correct region......The inherent demand for unbiasedness for some stereological estimators imposes a demand of not only positional uniform randomness but also isotropic randomness, i.e. directional uniform randomness. In order to comply with isotropy, one must perform a random rotation of the object of interest before...... is obeyed by randomizing the orientation of the virtual probe itself within the thick section. Overall, the benefit is that positional information is kept for any block and section of the specimen. As the Spatial Rotator is a 3D probe, data must be gathered from sections thicker than 25 micro meters to form...

  8. Rotational spectrum of tryptophan

    Energy Technology Data Exchange (ETDEWEB)

    Sanz, M. Eugenia, E-mail: maria.sanz@kcl.ac.uk; Cabezas, Carlos, E-mail: ccabezas@qf.uva.es; Mata, Santiago, E-mail: santiago.mata@uva.es; Alonso, Josè L., E-mail: jlalonso@qf.uva.es [Grupo de Espectroscopia Molecular (GEM), Edificio Quifima, Laboratorios de Espectroscopia y Bioespectroscopia, Unidad Asociada CSIC, Parque Científico Uva, Universidad de Valladolid, 47011 Valladolid (Spain)

    2014-05-28

    The rotational spectrum of the natural amino acid tryptophan has been observed for the first time using a combination of laser ablation, molecular beams, and Fourier transform microwave spectroscopy. Independent analysis of the rotational spectra of individual conformers has conducted to a definitive identification of two different conformers of tryptophan, with one of the observed conformers never reported before. The analysis of the {sup 14}N nuclear quadrupole coupling constants is of particular significance since it allows discrimination between structures, thus providing structural information on the orientation of the amino group. Both observed conformers are stabilized by an O–H···N hydrogen bond in the side chain and a N–H···π interaction forming a chain that reinforce the strength of hydrogen bonds through cooperative effects.

  9. Rotator cuff tendon connections with the rotator cable.

    Science.gov (United States)

    Rahu, Madis; Kolts, Ivo; Põldoja, Elle; Kask, Kristo

    2017-07-01

    The literature currently contains no descriptions of the rotator cuff tendons, which also describes in relation to the presence and characteristics of the rotator cable (anatomically known as the ligamentum semicirculare humeri). The aim of the current study was to elucidate the detailed anatomy of the rotator cuff tendons in association with the rotator cable. Anatomic dissection was performed on 21 fresh-frozen shoulder specimens with an average age of 68 years. The rotator cuff tendons were dissected from each other and from the glenohumeral joint capsule, and the superior glenohumeral, coracohumeral, coracoglenoidal and semicircular (rotator cable) ligaments were dissected. Dissection was performed layer by layer and from the bursal side to the joint. All ligaments and tendons were dissected in fine detail. The rotator cable was found in all specimens. It was tightly connected to the supraspinatus (SSP) tendon, which was partly covered by the infraspinatus (ISP) tendon. The posterior insertion area of the rotator cable was located in the region between the middle and inferior facets of the greater tubercle of the humerus insertion areas for the teres minor (TM), and ISP tendons were also present and fibres from the SSP extended through the rotator cable to those areas. The connection between the rotator cable and rotator cuff tendons is tight and confirms the suspension bridge theory for rotator cuff tears in most areas between the SSP tendons and rotator cable. In its posterior insertion area, the rotator cable is a connecting structure between the TM, ISP and SSP tendons. These findings might explain why some patients with relatively large rotator cuff tears can maintain seamless shoulder function.

  10. Rotator Cuff Exercises

    Science.gov (United States)

    ... Kids and Teens Pregnancy and Childbirth Women Men Seniors Your Health Resources Healthcare Management End-of-Life Issues Insurance & Bills Self Care Working With Your Doctor Drugs, Procedures & Devices Over-the- ...

  11. Assessment of hybrid rotation-translation scan schemes for in vivo animal SPECT imaging

    International Nuclear Information System (INIS)

    Xia Yan; Liu Yaqiang; Wang Shi; Ma Tianyu; Yao Rutao; Deng Xiao

    2013-01-01

    To perform in vivo animal single photon emission computed tomography imaging on a stationary detector gantry, we introduced a hybrid rotation-translation (HRT) tomographic scan, a combination of translational and limited angle rotational movements of the image object, to minimize gravity-induced animal motion. To quantitatively assess the performance of ten HRT scan schemes and the conventional rotation-only scan scheme, two simulated phantoms were first scanned with each scheme to derive the corresponding image resolution (IR) in the image field of view. The IR results of all the scan schemes were visually assessed and compared with corresponding outputs of four scan scheme evaluation indices, i.e. sampling completeness (SC), sensitivity (S), conventional system resolution (SR), and a newly devised directional spatial resolution (DR) that measures the resolution in any specified orientation. A representative HRT scheme was tested with an experimental phantom study. Eight of the ten HRT scan schemes evaluated achieved a superior performance compared to two other HRT schemes and the rotation-only scheme in terms of phantom image resolution. The same eight HRT scan schemes also achieved equivalent or better performance in terms of the four quantitative indices than the conventional rotation-only scheme. As compared to the conventional index SR, the new index DR appears to be a more relevant indicator of system resolution performance. The experimental phantom image obtained from the selected HRT scheme was satisfactory. We conclude that it is feasible to perform in vivo animal imaging with a HRT scan scheme and SC and DR are useful predictors for quantitatively assessing the performance of a scan scheme. (paper)

  12. The Nano-X Linear Accelerator: A Compact and Economical Cancer Radiotherapy System Incorporating Patient Rotation.

    Science.gov (United States)

    Eslick, Enid M; Keall, Paul J

    2015-10-01

    Rapid technological improvements in radiotherapy delivery results in improved outcomes to patients, yet current commercial systems with these technologies on board are costly. The aim of this study was to develop a state-of-the-art cancer radiotherapy system that is economical and space efficient fitting with current world demands. The Nano-X system is a compact design that is light weight combining a patient rotation system with a vertical 6 MV fixed beam. In this paper, we present the Nano-X system design configuration, an estimate of the system dimensions and its potential impact on shielding cost reductions. We provide an assessment of implementing such a radiotherapy system clinically, its advantages and disadvantages compared to a compact conventional gantry rotating linac. The Nano-X system has several differentiating features from current radiotherapy systems, it is [1] compact and therefore can fit into small vaults, [2] light weight, and [3] engineering efficient, i.e., it rotates a relatively light component and the main treatment delivery components are not under rotation (e.g., DMLCs). All these features can have an impact on reducing the costs of the system. In terms of shielding requirements, leakage radiation was found to be the dominant contributor to the Nano-X vault and as such no primary shielding was necessary. For a low leakage design, the Nano-X vault footprint and concrete volume required is 17 m2 and 35 m3 respectively, compared to 54 m2 and 102 m3 for a conventional compact linac vault, resulting in decreased costs in shielding. Key issues to be investigated in future work are the possible patient comfort concerns associated with the patient rotation system, as well as the magnitude of deformation and subsequent adaptation requirements. © The Author(s) 2014.

  13. The rotational spectrum of IBr

    International Nuclear Information System (INIS)

    Tiemann, E.; Moeller, T.

    1975-01-01

    The microwave spectrum of IBr was measured in the low rotational transition J = 3 → 2 in order to resolve the hyperfine structure as completely as possible. Rotational constants and quadrupole coupling constants were derived for both nuclei. The observation of the rotational spectrum in different vibrational states yields the vibrational dependence of the rotational constants as well as of the hyperfine parameters. The Dunham potential coefficients α 0 , α 1 , α 2 , α 3 are given. (orig.) [de

  14. Snakes and spin rotators

    International Nuclear Information System (INIS)

    Lee, S.Y.

    1990-01-01

    The generalized snake configuration offers advantages of either shorter total snake length and smaller orbit displacement in the compact configuration or the multi-functions in the split configuration. We found that the compact configuration can save about 10% of the total length of a snake. On other hand, the spilt snake configuration can be used both as a snake and as a spin rotator for the helicity state. Using the orbit compensation dipoles, the spilt snake configuration can be located at any distance on both sides of the interaction point of a collider provided that there is no net dipole rotation between two halves of the snake. The generalized configuration is then applied to the partial snake excitation. Simple formula have been obtained to understand the behavior of the partial snake. Similar principle can also be applied to the spin rotators. We also estimate the possible snake imperfections are due to various construction errors of the dipole magnets. Accuracy of field error of better than 10 -4 will be significant. 2 refs., 5 figs

  15. Underwater running device

    International Nuclear Information System (INIS)

    Kogure, Sumio; Matsuo, Takashiro; Yoshida, Yoji

    1996-01-01

    An underwater running device for an underwater inspection device for detecting inner surfaces of a reactor or a water vessel has an outer frame and an inner frame, and both of them are connected slidably by an air cylinder and connected rotatably by a shaft. The outer frame has four outer frame legs, and each of the outer frame legs is equipped with a sucker at the top end. The inner frame has four inner frame legs each equipped with a sucker at the top end. The outer frame legs and the inner frame legs are each connected with the outer frame and the inner frame by the air cylinder. The outer and the inner frame legs can be elevated or lowered (or extended or contracted) by the air cylinder. The sucker is connected with a jet pump-type negative pressure generator. The device can run and move by repeating attraction and releasing of the outer frame legs and the inner frame legs alternately while maintaining the posture of the inspection device stably. (I.N.)

  16. Numerical studies of Siberian snakes and spin rotators for RHIC

    International Nuclear Information System (INIS)

    Luccio, A.

    1995-01-01

    For the program of polarized protons in RHIC, two Siberian snakes and four spin rotators per ring will be used. The Snakes will produce a complete spin flip. Spin Rotators, in pairs, will rotate the spin from the vertical direction to the horizontal plane at a given insertion, and back to the vertical after the insertion. Snakes, 180 degrees apart and with their axis of spin precession at 90 degrees to each other, are an effective means to avoid depolarization of the proton beam in traversing resonances. Classical snakes and rotators are made with magnetic solenoids or with a sequence of magnetic dipoles with fields alternately directed in the radial and vertical direction. Another possibility is to use helical magnets, essentially twisted dipoles, in which the field, transverse the axis of the magnet, continuously rotates as the particles proceed along it. After some comparative studies, the authors decided to adopt for RHIC an elegant solution with four helical magnets both for the snakes and the rotators proposed by Shatunov and Ptitsin. In order to simplify the construction of the magnets and to minimize cost, four identical super conducting helical modules will be used for each device. Snakes will be built with four right-handed helices. Spin rotators with two right-handed and two left-handed helices. The maximum field will be limited to 4 Tesla. While small bore helical undulators have been built for free electron lasers, large super conducting helical magnets have not been built yet. In spite of this difficulty, this choice is dictated by some distinctive advantages of helical over more conventional transverse snakes/rotators: (i) the devices are modular, they can be built with arrangements of identical modules, (ii) the maximum orbit excursion in the magnet is smaller, (iii) orbit excursion is independent from the separation between adjacent magnets, (iv) they allow an easier control of the spin rotation and the orientation of the spin precession axis

  17. Fusion devices

    International Nuclear Information System (INIS)

    Fowler, T.K.

    1977-01-01

    Three types of thermonuclear fusion devices currently under development are reviewed for an electric utilities management audience. Overall design features of laser fusion, tokamak, and magnetic mirror type reactors are described and illustrated. Thrusts and trends in current research on these devices that promise to improve performance are briefly reviewed. Twenty photographs and drawings are included

  18. BRAKE DEVICE

    Science.gov (United States)

    O'Donnell, T.J.

    1959-03-10

    A brake device is described for utilization in connection with a control rod. The device comprises a pair of parallelogram link mechanisms, a control rod moveable rectilinearly therebetween in opposite directions, and shoes resiliently supported by the mechanism for frictional engagement with the control rod.

  19. CISM Course on Rotating Fluids

    CERN Document Server

    1992-01-01

    The volume presents a comprehensive overview of rotation effects on fluid behavior, emphasizing non-linear processes. The subject is introduced by giving a range of examples of rotating fluids encountered in geophysics and engineering. This is then followed by a discussion of the relevant scales and parameters of rotating flow, and an introduction to geostrophic balance and vorticity concepts. There are few books on rotating fluids and this volume is, therefore, a welcome addition. It is the first volume which contains a unified view of turbulence in rotating fluids, instability and vortex dynamics. Some aspects of wave motions covered here are not found elsewhere.

  20. COMMISSIONING SPIN ROTATORS IN RHIC

    International Nuclear Information System (INIS)

    MACKAY, W.W.; AHRENS, L.; BAI, M.; COURANT, E.D.; FISCHER, W.; HUANG, H.; LUCCIO, A.; MONTAG, C.; PILAT, F.; PTITSYN, V.; ROSER, T.; SATOGATA, T.; TRBOJEVIC, D.; VANZIEJTS, J.

    2003-01-01

    During the summer of 2002, eight superconducting helical spin rotators were installed into RHIC in order to control the polarization directions independently at the STAR and PHENIX experiments. Without the rotators, the orientation of polarization at the interaction points would only be vertical. With four rotators around each of the two experiments, we can rotate either or both beams from vertical into the horizontal plane through the interaction region and then back to vertical on the other side. This allows independent control for each beam with vertical, longitudinal, or radial polarization at the experiment. In this paper, we present results from the first run using the new spin rotators at PHENIX

  1. A mechanical rotator for neutron scattering measurements

    International Nuclear Information System (INIS)

    Thaler, A.; Northen, E.; Aczel, A. A.; MacDougall, G. J.

    2016-01-01

    We have designed and built a mechanical rotation system for use in single crystal neutron scattering experiments at low temperatures. The main motivation for this device is to facilitate the application of magnetic fields transverse to a primary training axis, using only a vertical cryomagnet. Development was done in the context of a triple-axis neutron spectrometer, but the design is such that it can be generalized to a number of different instruments or measurement techniques. Here, we discuss some of the experimental constraints motivating the design, followed by design specifics, preliminary experimental results, and a discussion of potential uses and future extension possibilities.

  2. NMSBA - Twist Resist - Rotational Exercise Module

    Energy Technology Data Exchange (ETDEWEB)

    Walker, Aaron [Twist Resist, Albuquerque, NM (United States); Reece, Blake D. [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Berger, Jason E. [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Guido, Steven Frank [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Linker, Taylor [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)

    2016-08-01

    This report contains a summary of the work completed to develop a modular, rotational exercise device. In the report are images, diagrams, and explanations of the efforts contributed to the project since its inception. The purpose of this document is to provide a walk-through of the progress on this project, from the initial design concepts to the final design and work done, so that the customer (Twist Resist), or individuals/firms who work on this project in the future will have a springboard of ideas/concepts to work from.

  3. Cooling device of superconducting coils

    International Nuclear Information System (INIS)

    Duthil, R.; Lottin, J.C.

    1985-01-01

    This device is rotating around an horizontal axis. The superconducting coils are contained in a cryogenic enclosure feeded in liquid helium forced circulation. They are related to an electric generator by electric mains each of them comprising a gas exchanger, and an exchanger-evaporator set between the cryogenic device and those exchangers. The exchanger-evaporator is aimed at dissipating the heat arriving by conductors connected to the superconducting coils. According to the invention, the invention includes an annular canalization with horizontal axis in which the connection conductors bathe in liquid helium [fr

  4. Design Concepts for a Continuously Rotating Active Magnetic Regenerator

    DEFF Research Database (Denmark)

    Bahl, Christian Robert Haffenden; Engelbrecht, Kurt; Bjørk, Rasmus

    2010-01-01

    Design considerations for a prototype magnetic refrigeration device with a continuously rotating AMR are presented. Building the AMR from stacks of elongated plates of the perovskite oxide material La0.67Ca0.33-xSrxMn1.05O3, gives both a low pressure drop and allows grading of the Curie temperatu...

  5. CONFERENCE: Muon spin rotation

    Energy Technology Data Exchange (ETDEWEB)

    Karlsson, Erik

    1986-11-15

    An international physics conference centred on muons without a word about leptons, weak interactions, EMC effects, exotic decay modes or any other standard high energy physics jargon. Could such a thing even have been imagined ten years ago? Yet about 120 physicists and chemists from 16 nations gathered at the end of June in Uppsala (Sweden) for their fourth meeting on Muon Spin Rotation, Relaxation and Resonance, without worrying about the muon as an elementary particle. This reflects how the experimental techniques based on the muon spin interactions have reached maturity and are widely recognized by condensed matter physicists and specialized chemists as useful tools.

  6. Autonomous quantum rotator

    DEFF Research Database (Denmark)

    Fogedby, Hans C.; Imparato, Alberto

    2018-01-01

    to a directed rotary motion. At variance with the classical case, the thermal fluctuations in the baths give rise to a non-vanishing average torque contribution; this is a genuine quantum effect akin to the Casimir effect. In the steady state the heat current flowing between the two baths is systematically......, the rotator cannot work either as a heat pump or as a heat engine. We finally use our exact results to extend an ab initio quantum simulation algorithm to the out-of-equilibrium regime. Copyright (C) EPLA, 2018...

  7. Rotating specimen rack repair

    International Nuclear Information System (INIS)

    Miller, G.E.; Rogers, P.J.; Nabor, W.G.; Bair, H.

    1984-01-01

    In 1980, an operator at the UCI TRIGA Reactor noticed difficulties with the rotation of the specimen rack. Investigations showed that the drive bearing in the rack had failed and allowed the bearings to enter the rack. After some time of operation in static mode it was decided that installation of a bearing substitute - a graphite sleeve - would be undertaken. Procedures were written and approved for removal of the rack, fabrication and installation of the sleeve, and re-installation of the rack. This paper describes these procedures in some detail. Detailed drawings of the necessary parts may be obtained from the authors

  8. Rotational anomalies without anyons

    International Nuclear Information System (INIS)

    Hagen, C.R.

    1985-01-01

    A specific field theory is proposed in two spatial dimensions which has anomalous rotational properties. Although this might be expected to lead to a concrete realization of what Wilczek refers to as the anyon, it is shown by utilizing the transformation properties of the system and the statistics of the underlying charge fields that anyonic interpolations between bosons and fermions do not occur. This leads to the suggestion that anyons inferred from semiclassical considerations will not survive the transition to a fully relativistic field theory

  9. Rotating electrical machines

    CERN Document Server

    Le Doeuff, René

    2013-01-01

    In this book a general matrix-based approach to modeling electrical machines is promulgated. The model uses instantaneous quantities for key variables and enables the user to easily take into account associations between rotating machines and static converters (such as in variable speed drives).   General equations of electromechanical energy conversion are established early in the treatment of the topic and then applied to synchronous, induction and DC machines. The primary characteristics of these machines are established for steady state behavior as well as for variable speed scenarios. I

  10. Advanced Small Animal Conformal Radiation Therapy Device.

    Science.gov (United States)

    Sharma, Sunil; Narayanasamy, Ganesh; Przybyla, Beata; Webber, Jessica; Boerma, Marjan; Clarkson, Richard; Moros, Eduardo G; Corry, Peter M; Griffin, Robert J

    2017-02-01

    We have developed a small animal conformal radiation therapy device that provides a degree of geometrical/anatomical targeting comparable to what is achievable in a commercial animal irradiator. small animal conformal radiation therapy device is capable of producing precise and accurate conformal delivery of radiation to target as well as for imaging small animals. The small animal conformal radiation therapy device uses an X-ray tube, a robotic animal position system, and a digital imager. The system is in a steel enclosure with adequate lead shielding following National Council on Radiation Protection and Measurements 49 guidelines and verified with Geiger-Mueller survey meter. The X-ray source is calibrated following AAPM TG-61 specifications and mounted at 101.6 cm from the floor, which is a primary barrier. The X-ray tube is mounted on a custom-made "gantry" and has a special collimating assembly system that allows field size between 0.5 mm and 20 cm at isocenter. Three-dimensional imaging can be performed to aid target localization using the same X-ray source at custom settings and an in-house reconstruction software. The small animal conformal radiation therapy device thus provides an excellent integrated system to promote translational research in radiation oncology in an academic laboratory. The purpose of this article is to review shielding and dosimetric measurement and highlight a few successful studies that have been performed to date with our system. In addition, an example of new data from an in vivo rat model of breast cancer is presented in which spatially fractionated radiation alone and in combination with thermal ablation was applied and the therapeutic benefit examined.

  11. Rotational Acceleration during Head Impact Resulting from Different Judo Throwing Techniques

    OpenAIRE

    MURAYAMA, Haruo; HITOSUGI, Masahito; MOTOZAWA, Yasuki; OGINO, Masahiro; KOYAMA, Katsuhiro

    2014-01-01

    Most severe head injuries in judo are reported as acute subdural hematoma. It is thus necessary to examine the rotational acceleration of the head to clarify the mechanism of head injuries. We determined the rotational acceleration of the head when the subject is thrown by judo techniques. One Japanese male judo expert threw an anthropomorphic test device using two throwing techniques, Osoto-gari and Ouchigari. Rotational and translational head accelerations were measured with and without an ...

  12. Assessing the quality of proton PBS treatment delivery using machine log files: comprehensive analysis of clinical treatments delivered at PSI Gantry 2

    International Nuclear Information System (INIS)

    Scandurra, D; Albertini, F; Van der Meer, R; Meier, G; Weber, D C; Bolsi, A; Lomax, A

    2016-01-01

    Pencil beam scanning (PBS) proton therapy requires the delivery of many thousand proton beams, each modulated for position, energy and monitor units, to provide a highly conformal patient treatment. The quality of the treatment is dependent on the delivery accuracy of each beam and at each fraction. In this work we describe the use of treatment log files, which are a record of the machine parameters for a given field delivery on a given fraction, to investigate the integrity of treatment delivery compared to the nominal planned dose. The dosimetry-relevant log file parameters are used to reconstruct the 3D dose distribution on the patient anatomy, using a TPS-independent dose calculation system. The analysis was performed for patients treated at Paul Scherrer Institute on Gantry 2, both for individual fields and per series (or plan), and delivery quality was assessed by determining the percentage of voxels in the log file dose distribution within  +/−  1% of the nominal dose. It was seen that, for all series delivered, the mean pass rate is 96.4%. Furthermore, this work establishes a correlation between the delivery quality of a field and the beam position accuracy. This correlation is evident for all delivered fields regardless of individual patient or plan characteristics. We have also detailed further usefulness of log file analysis within our clinical workflow. In summary, we have highlighted that the integrity of PBS treatment delivery is dependent on daily machine performance and is specifically highly correlated with the accuracy of beam position. We believe this information will be useful for driving machine performance improvements in the PBS field. (paper)

  13. Experimental and numerical results of a high frequency rotating active magnetic refrigerator

    DEFF Research Database (Denmark)

    Lozano, Jaime; Engelbrecht, Kurt; Bahl, Christian

    2012-01-01

    Experimental results for a recently developed prototype magnetic refrigeration device at The Technical University of Denmark (DTU) were obtained and compared with numerical simulation results. A continuously rotating active magnetic regenerator (AMR) using 2.8 kg packed sphere regenerators...

  14. Experimental and numerical results of a high frequency rotating active magnetic refrigerator

    DEFF Research Database (Denmark)

    Lozano, Jaime; Engelbrecht, Kurt; Bahl, Christian R.H.

    2014-01-01

    Experimental results for a recently developed prototype magnetic refrigeration device at the Technical University of Denmark (DTU) were obtained and compared with numerical simulation results. A continuously rotating active magnetic regenerator (AMR) using 2.8 kg packed sphere regenerators...

  15. Improved thermal monitoring of rotating machine insulation

    International Nuclear Information System (INIS)

    Stone, G.C.; Sedding, H.G.; Bernstein, B.S.

    1991-01-01

    Aging of motor and generator insulation is most often induced as a result of operation at high temperatures. In spite of this knowledge, stator and rotor temperatures are only crudely monitored in existing machines. In EPRI project RP2577-1, three new means of detecting machine temperatures were successfully developed. Two of the techniques, the Electronic Rotor Temperature Sensor and the Passive Rotor Temperature Sensor, were specifically developed to give point temperature readings on turbine generator rotor windings. The Insulation Sniffer allows operators to determine when any electrical insulation in a motor is overheating. Another electronic device, called the Thermal Life Indicator, helps operators and maintenance personnel determine how accumulated operation has affected the remaining life of the insulation in rotating machines. These new devices permit nuclear station operators to avoid hazardous operating conditions and will help to determine priorities for maintenance and plant life extension programs

  16. SU-F-T-649: Dosimetric Evaluation of Non-Coplanar Arc Therapy Using a Novel Rotating Gamma Ray System

    Energy Technology Data Exchange (ETDEWEB)

    Eldib, A; Chibani, O; Jin, L; Fan, J; Veltchev, I; Ma, C [Fox Chase Cancer Center, Philadelphia, PA (United States); Mora, G [Universidade de Lisboa, Codex, Lisboa (Portugal); Li, J [Cyber Medical Inc, Xian, Shaanxi (China)

    2016-06-15

    Purpose: Stereotactic intra and extra-cranial body radiation therapy has evolved with advances in treatment accuracy, effective radiation dose, and parameters necessary to maximize machine capabilities. Novel gamma systems with a ring type gantry were developed having the ability to perform oblique arcs. The aim of this study is to explore the dosimetric advantages of this new system. Methods: The rotating Gamma system is named CybeRay (Cyber Medical Corp., Xian, China). It has a treatment head of 16 cobalt-60 sources focused to the isocenter, which can rotate 360° on the ring gantry and swing 35° in the superior direction. Treatment plans were generated utilizing our in-house Monte Carlo treatment planning system. A cylindrical phantom was modeled with 2mm voxel size. Dose inside the cylindrical phantom was calculated for coplanar and non-coplanar arcs. Dosimetric differences between CybeRay cobalt beams and CyberKnife 6MV beams were compared in a lung phantom and for previously treated SBRT patients. Results: The full width at half maxima of cross profiles in the S-I direction for the coplanar setup matched the cone sizes, while for the non-coplanar setup, FWHM was larger by 2mm for a 10mm cone and about 5mm for larger cones. In the coronal and sagittal view, coplanar beams showed elliptical shaped isodose lines, while non-coplanar beams showed circular isodose lines. Thus proper selection of the oblique angle and cone size can aid optimal dose matching to the target volume. Comparing a single 5mm cone from CybeRay to that from CyberKnife showed similar penumbra in a lung phantom but CybeRay had significant lower doses beyond lung tissues. Comparable treatment plans were obtained with CybeRay as that from CyberKnife.ConclusionThe noncoplanar multiple source arrangement of CybeRay will be of great clinical benefits for stereotactic intra and extra-cranial radiation therapy.

  17. Electromagnetic brake/clutch device

    Science.gov (United States)

    Vranish, John M. (Inventor)

    1994-01-01

    An electromagnetic brake/clutch device includes a drive shaft supported by at least one bearing for transmitting torque, a housing, affixed to prevent its rotation, surrounding the drive shaft, and an electromagnetically activated device within the housing to selectively prevent and allow rotation of the drive shaft. The electromagnetically activated device includes a plurality of cammed rollers to prevent counter-clockwise rotation of the drive shaft. The drive shaft includes a circumferential disk and the housing includes a reaction ring for engagement with the plurality of cammed rollers. The plurality of cammed rollers are released from engagement with the circumferential disk and the reaction ring by a plurality of tripping mechanisms within the housing. The tripping action uses the locking force to act as a release force merely by changing the boundary conditions of the roller interface angles. The tripping mechanisms include trippers for disengaging the plurality of cammed rollers and an anvil shaped portion for providing lateral movement of the trippers. The plurality of cammed rollers is preloaded to engagement with the circumferential disk and reaction ring by a spring, and is located with respect to an adjacent tripping mechanism with another spring.

  18. Sealing devices

    International Nuclear Information System (INIS)

    Coulson, R.A.

    1980-01-01

    A sealing device for minimising the leakage of toxic or radioactive contaminated environments through a biological shield along an opening through which a flexible component moves that penetrates the shield. The sealing device comprises an outer tubular member which extends over a length not less than the maximum longitudinal movement of the component along the opening. An inner sealing block is located intermediate the length of the component by connectors and is positioned in the bore of the outer tubular member to slide in the bore and effect a seal over the entire longitudinal movement of the component. The cross-section of the device may be circular and the block may be of polytetrafluoroethylene or of nylon impregnated with molybdenum or may be metallic. A number of the sealing devices may be combined into an assembly for a plurality of adjacent longitudinally movable components, each adapted to sustain a tensile load, providing the various drives of a master-slave manipulator. (author)

  19. Global low-frequency modes in weakly ionized magnetized plasmas: effects of equilibrium plasma rotation

    International Nuclear Information System (INIS)

    Sosenko, P.; Pierre, Th.; Zagorodny, A.

    2004-01-01

    The linear and non-linear properties of global low-frequency oscillations in cylindrical weakly ionized magnetized plasmas are investigated analytically for the conditions of equilibrium plasma rotation. The theoretical results are compared with the experimental observations of rotating plasmas in laboratory devices, such as Mistral and Mirabelle in France, and KIWI in Germany. (authors)

  20. Ferroelectric devices

    CERN Document Server

    Uchino, Kenji

    2009-01-01

    Updating its bestselling predecessor, Ferroelectric Devices, Second Edition assesses the last decade of developments-and setbacks-in the commercialization of ferroelectricity. Field pioneer and esteemed author Uchino provides insight into why this relatively nascent and interdisciplinary process has failed so far without a systematic accumulation of fundamental knowledge regarding materials and device development.Filling the informational void, this collection of information reviews state-of-the-art research and development trends reflecting nano and optical technologies, environmental regulat

  1. Rotation of hard particles in a soft matrix

    Science.gov (United States)

    Yang, Weizhu; Liu, Qingchang; Yue, Zhufeng; Li, Xiaodong; Xu, Baoxing

    Soft-hard materials integration is ubiquitous in biological materials and structures in nature and has also attracted growing attention in the bio-inspired design of advanced functional materials, structures and devices. Due to the distinct difference in their mechanical properties, the rotation of hard phases in soft matrixes upon deformation has been acknowledged, yet is lack of theory in mechanics. In this work, we propose a theoretical mechanics framework that can describe the rotation of hard particles in a soft matrix. The rotation of multiple arbitrarily shaped, located and oriented particles with perfectly bonded interfaces in an elastic soft matrix subjected to a far-field tensile loading is established and analytical solutions are derived by using complex potentials and conformal mapping methods. Strong couplings and competitions of the rotation of hard particles among each other are discussed by investigating numbers, relative locations and orientations of particles in the matrix at different loading directions. Extensive finite element analyses are performed to validate theoretical solutions and good agreement of both rotation and stress field between them are achieved. Possible extensions of the present theory to non-rigid particles, viscoelastic matrix and imperfect bonding are also discussed. Finally, by taking advantage of the rotation of hard particles, we exemplify an application in a conceptual design of soft-hard material integrated phononic crystal and demonstrate that phononic band gaps can be successfully tuned with a high accuracy through the mechanical tension-induced rotation of hard particles. The present theory established herein is expected to be of immediate interests to the design of soft-hard materials integration based functional materials, structures and devices with tunable performance via mechanical rotation of hard phases.

  2. Performance test of wet type decontamination device

    International Nuclear Information System (INIS)

    Lee, E. P.; Kim, E. G.; Min, D. K.; Jun, Y. B.; Lee, H. K.; Seu, H. S.; Kwon, H. M.; Hong, K.P.

    2003-01-01

    The intervention area located at rear hot cell can be contaminated by hot cell maintenance work. For effective decontamination of the intervention floor a wet type decontamination device was developed. The device was assembled with a brush rotating part, a washing liquid supplying part, an intake part for recovering contaminated liquid and a device moving cart part. The device was made of stainless steel for easy decontamination and corrosion resistance. The function test carried out at intervention area of the PIE facility showed good performance

  3. A low frequency rotational energy harvesting system

    International Nuclear Information System (INIS)

    Febbo, M; Machado, S P; Ramirez, J M; Gatti, C D

    2016-01-01

    This paper presents a rotary power scavenging unit comprised of two systems of flexible beams connected by two masses which are joined by means of a spring, considering a PZT (QP16N, Midé Corporation) piezoelectric sheet mounted on one of the beams. The energy harvesting (EH) system is mounted rigidly on a rotating hub. The gravitational force on the masses causes sustained oscillatory motion in the flexible beams as long as there is rotary motion. The intention is to use the EH system in the wireless autonomous monitoring of wind turbines under different wind conditions. Specifically, the development is oriented to monitor the dynamic state of the blades of a wind generator of 30 KW which rotates between 50 and 150 rpm. The paper shows a complete set of experimental results on three devices, modifying the amount of beams in the frame supporting the system. The results show an acceptable sustained voltage generation for the expected range, in the three proposed cases. Therefore, it is possible to use this system for generating energy in a low-frequency rotating environment. As an alternative, the system can be easily adapted to include an array of piezoelectric sheets to each of the beams, to provide more power generation. (paper)

  4. An electronic pan/tilt/magnify and rotate camera system

    International Nuclear Information System (INIS)

    Zimmermann, S.; Martin, H.L.

    1992-01-01

    A new camera system has been developed for omnidirectional image-viewing applications that provides pan, tilt, magnify, and rotational orientation within a hemispherical field of view (FOV) without any moving parts. The imaging device is based on the fact that the image from a fish-eye lens, which produces a circular image of an entire hemispherical FOV, can be mathematically corrected using high-speed electronic circuitry. More specifically, an incoming fish-eye image from any image acquisition source is captured in the memory of the device, a transformation is performed for the viewing region of interest and viewing direction, and a corrected image is output as a video image signal for viewing, recording, or analysis. The image transformation device can provide corrected images at frame rates compatible with RS-170 standard video equipment. As a result, this device can accomplish the functions of pan, tilt, rotation, and magnification throughout a hemispherical FOV without the need for any mechanical devices. Multiple images, each with different image magnifications and pan-tilt-rotate parameters, can be obtained from a single camera

  5. Lunar Rotation, Orientation and Science

    Science.gov (United States)

    Williams, J. G.; Ratcliff, J. T.; Boggs, D. H.

    2004-12-01

    The Moon is the most familiar example of the many satellites that exhibit synchronous rotation. For the Moon there is Lunar Laser Ranging measurements of tides and three-dimensional rotation variations plus supporting theoretical understanding of both effects. Compared to uniform rotation and precession the lunar rotational variations are up to 1 km, while tidal variations are about 0.1 m. Analysis of the lunar variations in pole direction and rotation about the pole gives moment of inertia differences, third-degree gravity harmonics, tidal Love number k2, tidal dissipation Q vs. frequency, dissipation at the fluid-core/solid-mantle boundary, and emerging evidence for an oblate boundary. The last two indicate a fluid core, but a solid inner core is not ruled out. Four retroreflectors provide very accurate positions on the Moon. The experience with the Moon is a starting point for exploring the tides, rotation and orientation of the other synchronous bodies of the solar system.

  6. A compact rotating dilution refrigerator

    Science.gov (United States)

    Fear, M. J.; Walmsley, P. M.; Chorlton, D. A.; Zmeev, D. E.; Gillott, S. J.; Sellers, M. C.; Richardson, P. P.; Agrawal, H.; Batey, G.; Golov, A. I.

    2013-10-01

    We describe the design and performance of a new rotating dilution refrigerator that will primarily be used for investigating the dynamics of quantized vortices in superfluid 4He. All equipment required to operate the refrigerator and perform experimental measurements is mounted on two synchronously driven, but mechanically decoupled, rotating carousels. The design allows for relative simplicity of operation and maintenance and occupies a minimal amount of space in the laboratory. Only two connections between the laboratory and rotating frames are required for the transmission of electrical power and helium gas recovery. Measurements on the stability of rotation show that rotation is smooth to around 10-3 rad s-1 up to angular velocities in excess of 2.5 rad s-1. The behavior of a high-Q mechanical resonator during rapid changes in rotation has also been investigated.

  7. Wormholes immersed in rotating matter

    Directory of Open Access Journals (Sweden)

    Christian Hoffmann

    2018-03-01

    Full Text Available We demonstrate that rotating matter sets the throat of an Ellis wormhole into rotation, allowing for wormholes which possess full reflection symmetry with respect to the two asymptotically flat spacetime regions. We analyze the properties of this new type of rotating wormholes and show that the wormhole geometry can change from a single throat to a double throat configuration. We further discuss the ergoregions and the lightring structure of these wormholes.

  8. Internal rotation of the Sun

    International Nuclear Information System (INIS)

    Duvall, T.L. Jr.; Goode, P.R.; Gouch, D.O.

    1984-01-01

    The frequency difference between prograde and retrograde sectoral solar oscillations is analysed to determine the rotation rate of the solar interior, assuming no latitudinal dependence. Much of the solar interior rotates slightly less rapidly than the surface, while the innermost part apparently rotates more rapidly. The resulting solar gravitational quadrupole moment is J 2 = (1.7 +- 0.4) x 10 -7 and provides a negligible contribution to current planetary tests of Einstein's theory of general relativity. (author)

  9. Earth's variable rotation

    Science.gov (United States)

    Hide, Raymond; Dickey, Jean O.

    1991-01-01

    Recent improvements in geodetic data and practical meteorology have advanced research on fluctuations in the earth's rotation. The interpretation of these fluctuations is inextricably linked with studies of the dynamics of the earth-moon system and dynamical processes in the liquid metallic core of the earth (where the geomagnetic field originates), other parts of the earth's interior, and the hydrosphere and atmosphere. Fluctuations in the length of the day occurring on decadal time scales have implications for the topographay of the core-mantle boundary and the electrical, magnetic, ande other properties of the core and lower mantle. Investigations of more rapid fluctuations bear on meteorological studies of interannual, seasonal, and intraseasonal variations in the general circulation of the atmosphere and the response of the oceans to such variations.

  10. Sporcularda rotator cuff problemleri

    OpenAIRE

    Guven, Osman; Guven, Zeynep; Gundes, Hakan; Yalcin, Selim

    2004-01-01

    Rotator cuff tendinitinin etyolojisinde genellikle birden çok faktörün kombinasyonu görülür. Yüzme, raket sporları ve fırlatma sporlarının özellikle gelişmiş ülkelerde giderek yaygınlaşması bu konuya olan ilginin artmasına sebep olmuştur. Eski konseptlerde aktif bir sporcuda tedavinin başarısı genellikle eski atletik seviyesine dönmesi ile ölçülürdü. Son zamanlarda atletik tekniklerin analizi, atroskopik evaluasyon gibi yeni bir Iükse sahip olmamız ve Iiteratürün yeniden gözden geçirilmesi il...

  11. Device for selectively securing an object to a shaft

    International Nuclear Information System (INIS)

    Calizano, F.; Chauvel, M.

    1984-01-01

    A magnetic tape reel is secured to a reel drive motor shaft by a device including a hub secured to the shaft, a plurality of shoes, and actuating means for thrusting the shoes against the reel and for releasing them from contact with the reel. The securing device includes a cam mounted on the shaft in combination with a locking device including the cam periphery. The locking device thrusts the shoes against the object and is operated by rotation of the motor. The cam is immobilized in rotation while the shoes are locked

  12. A device for transferring, in particular, small particles

    International Nuclear Information System (INIS)

    1974-01-01

    The invention relates to a transfer device, in particular for transferring small particles, comprising a helical channel made in the tube inner surface, a device for causing the tube to rotate about its longitudinal axis, a rotating joint adapted to close one of the tube extremities, a device for inserting a substance in the form of granules or of fluid particles into said tube through said joint, and a device for collecting and discharging said substance at the tube opposite end. This can applied to the transfer of small spherical particles e.g. of fuel [fr

  13. Laser working device

    International Nuclear Information System (INIS)

    Shibanuma, Kiyoshi; Kakudate, Satoshi; Oka, Kiyoshi; Terakado, Takuya; Kondo, Mitsunori; Munakata, Tadashi; Makino, Yoshinobu; Honda, Keizo.

    1995-01-01

    A transmission pipe transmits laser beams along an axis thereof, and is inserted at the top end to a pipeline to be fabricated. A flat mirror is secured to the top end of the transmission pipe, and laser beams are reflected by the mirror, passed through a fabrication nozzle and focused to a fabrication point in the pipeline to be fabricated. A lens-type light focusing system is guided to the fabrication point by a plurality of rollers rotatable in the axial direction disposed in circumferential direction each at an equal pitch at the outer circumference of the transmission pipe. A centering mechanism is disposed for keeping the transmission pipe coaxially with the pipeline to be fabricated. Further, there are also disposed a mirror-type light focusing optical system for focusing light by a paraboloidal mirror and a spherical vehicle rotatable in all directions. A laser fabrication device can be reduced in the size, and it can be used in a high temperature and highly radioactive circumstance. (N.H.)

  14. Device for guiding various diameter size cable

    Energy Technology Data Exchange (ETDEWEB)

    Litvinov, I.M.; Klauzer, L.P.; Yeganov, L.I.; Zaripov, A.M.

    1982-01-30

    A design is submitted for a device to guide cable of various diameters. This device consists of a profiled multiple-pass roller, and clamps for uniting cut cable. This design is simplified by allowing both the rollers and their supports to rotate on the roller axis thus facilitating interaction with the clamps. The working surface of the supports is evolute while the outside surfaces have screw-channels for transfering the cable from one roller pass to the other.

  15. A two-in-one Faraday rotator mirror exempt of active optical alignment.

    Science.gov (United States)

    Wan, Qiong; Wan, Zhujun; Liu, Hai; Liu, Deming

    2014-02-10

    A two-in-one Faraday rotator mirror was presented, which functions as two independent Faraday rotation mirrors with a single device. With the introduction of a reflection lens as substitution of the mirror in traditional structure, this device is characterized by exemption of active optical alignment for the designers and manufacturers of Faraday rotator mirrors. A sample was fabricated by passive mechanical assembly. The insertion loss was measured as 0.46 dB/0.50 dB for the two independent ports, respectively.

  16. SU-E-P-55: The Reaserch of Cervical Cancer Delivered with Constant Dose Rate and Gantry Speed Arc Therapy(CDR-CAS-IMAT) On Conventional Linac

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, R; Bai, W; Chi, Z; Gao, C; Xiaomei, F [The Fourth Hospital of Hebei Medical University, Shijiazhuang, Hebei (China); Gao, Y [Hebei General Hospital, Shijiazhuang, Hebei (China)

    2015-06-15

    Purpose: Postoperative cervical cancer patients with large target volume and the target shape is concave, treatmented with static intensity-modulated radiotherapy (IMRT) is time consuming. The purpose of this study is to investigate using constant dose rate and gantry speed arc therapy(CDR-CAS-IMAT) on conventional linear accelrator, by comparing with the IMRT technology to evaluate the performance of CDR-CAS-IMAT on postoperative cervical cancer patients. Methods: 18 cervical cancer patients treated with IMRT on Varian 23IX were replanted using CDR-CAS-IMAT. The plans were generated on Oncentra v4.1 planning system, PTV was prescribed to 50.4 Gy in 28 fractions. Plans were evaluated based on the ability to meet the dose volume histogram. The homogeneity index (HI), conformity index (CI) of target volume, the dose of organs at risk, radiation delivery time and monitor units were also compared. SPSS 19.0 software paired T-test analysis was carried out on the two sets of data. Results: Compared with the IMRT plans PTV’s CI (t= 3.85, P =0.001), CTV’s CI, HI, D90, D95, D98, V95, V98, V100 (t=4.21, −3.18, 2.13, 4.65, 7.79, 2.29, 6.00, 2.13, p=0.001, 0.005, 0.049, 0.000, 0.000, 0.035, 0.000, 0.049), and cord D2 and rectum V40 (t=−2.65, −2.47, p= P =0.017, 0.025), and treatment time and MU (t=−36.0, −6.26, P =0.000, 0.000) were better than that of IMRT group. But the IMRT plans in terms of decreasing bladder V50, bowel V30 (t=2.14, 3.00, P =0.048, 0.008) and low dose irradiation volume were superior to that of CDR-CAS-IMAT plans. There were no significant differences in other statistical index. Conclusion: Cervical cancer patients with CDR-CAS-IMAT on Varian Clinical 23IX can get equivalent or superior dose distribution compared with the IMRT technology. IMAT have much less treatment time and MU can reduce the uncertainty factor and patient discomfort in treatment. This work was supported by the Medical Science Foundation of the health department of Hebei

  17. Medical Devices

    NARCIS (Netherlands)

    Verkerke, Gijsbertus Jacob; Mahieu, H.F.; Geertsema, A.A.; Hermann, I.F.; van Horn, J.R.; Hummel, J. Marjan; van Loon, J.P.; Mihaylov, D.; van der Plaats, A.; Schraffordt Koops, H.; Schutte, H.K.; Veth, R.P.H.; de Vries, M.P.; Rakhorst, G.; Shi, Donglu

    2004-01-01

    The development of new medical devices is a very time-consuming and costly process. Besides the time between the initial idea and the time that manufacturing and testing of prototypes takes place, the time needed for the development of production facilities, production of test series, marketing,

  18. Electrochemical device

    Science.gov (United States)

    Grimes, Patrick G.; Einstein, Harry; Bellows, Richard J.

    1988-01-12

    A tunnel protected electrochemical device features channels fluidically communicating between manifold, tunnels and cells. The channels are designed to provide the most efficient use of auxiliary power. The channels have a greater hydraulic pressure drop and electrical resistance than the manifold. This will provide a design with the optimum auxiliary energy requirements.

  19. Assistive Devices

    Science.gov (United States)

    If you have a disability or injury, you may use a number of assistive devices. These are tools, products or types of equipment that help you perform tasks and activities. They may help you move around, see, communicate, eat, or get ...

  20. Detection device

    Science.gov (United States)

    Smith, J.E.

    1981-02-27

    The present invention is directed to a detection device comprising: (1) an entrance chamber; (2) a central chamber; and (3) an exit chamber. The central chamber includes an ionizing gas, anode, and means for connecting the anode with an external power supply and pulse counter.

  1. Rotation of the planet mercury.

    Science.gov (United States)

    Jefferys, W H

    1966-04-08

    The equations of motion for the rotation of Mercury are solved for the general case by an asymptotic expansion. The findings of Liu and O'Keefe, obtained by numerical integration of a special case, that it is possible for Mercury's rotation to be locked into a 2:3 resonance with its revolution, are confirmed in detail. The general solution has further applications.

  2. Stabilities of MHD rotational discontinuities

    International Nuclear Information System (INIS)

    Wang, S.

    1984-11-01

    In this paper, the stabilities of MHD rotational discontinuities are analyzed. The results show that the rotational discontinuities in an incompressible magnetofluid are not always stable with respect to infinitesimal perturbation. The instability condition in a special case is obtained. (author)

  3. Optical isolation by Faraday rotator

    International Nuclear Information System (INIS)

    Kasai, Takeshi; Matsushima, Isao; Nemoto, Fusashi; Yano, Masaaki

    1984-01-01

    Three Faraday rotators designed as optical isolators in a high power glass laser system are described. The spatial fluctuation of applied magnetic field is less than 1% throughout the Faraday glass rod. The Faraday rotators transmit more than 80% of the forward-going laser light and reject more than 96% of the backward-going light. (author)

  4. Ultra-thin, single-layer polarization rotator

    Directory of Open Access Journals (Sweden)

    T. V. Son

    2016-08-01

    Full Text Available We demonstrate light polarization control over a broad spectral range by a uniform layer of vanadium dioxide as it undergoes a phase transition from insulator to metal. Changes in refractive indices create unequal phase shifts on s- and p-polarization components of incident light, and rotation of linear polarization shows intensity modulation by a factor of 103 when transmitted through polarizers. This makes possible polarization rotation devices as thin as 50 nm that would be activated thermally, optically or electrically.

  5. Cooling device in thermonuclear device

    International Nuclear Information System (INIS)

    Honda, Tsutomu.

    1988-01-01

    Purpose: To prevent loss of cooling effect over the entire torus structure directly after accidental toubles in a cooling device of a thermonuclear device. Constitution: Coolant recycling means of a cooling device comprises two systems, which are alternately connected with in-flow pipeways and exit pipeways of adjacent modules. The modules are cooled by way of the in-flow pipeways and the exist pipeways connected to the respective modules by means of the coolant recycling means corresponding to the respective modules. So long as one of the coolant recycling means is kept operative, since every one other modules of the torus structure is still kept cooled, the heat generated from the module put therebetween, for which the coolant recycling is interrupted, is removed by means of heat conduction or radiation from the module for which the cooling is kept continued. No back-up emergency cooling system is required and it can provide high economic reliability. (Kamimura, M.)

  6. Rotational superradiance in fluid laboratories

    CERN Document Server

    Cardoso, Vitor; Richartz, Mauricio; Weinfurtner, Silke

    2016-01-01

    Rotational superradiance has been predicted theoretically decades ago, and is the chief responsible for a number of important effects and phenomenology in black hole physics. However, rotational superradiance has never been observed experimentally. Here, with the aim of probing superradiance in the lab, we investigate the behaviour of sound and surface waves in fluids resting in a circular basin at the center of which a rotating cylinder is placed. We show that with a suitable choice for the material of the cylinder, surface and sound waves are amplified. By confining the superradiant modes near the rotating cylinder, an instability sets in. Our findings are experimentally testable in existing fluid laboratories and hence offer experimental exploration and comparison of dynamical instabilities arising from rapidly rotating boundary layers in astrophysical as well as in fluid dynamical systems.

  7. Advances in Rotational Seismic Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Pierson, Robert [Applied Technology Associates, Albuquerque, NM (United States); Laughlin, Darren [Applied Technology Associates, Albuquerque, NM (United States); Brune, Robert [Applied Technology Associates, Albuquerque, NM (United States)

    2016-10-19

    Rotational motion is increasingly understood to be a significant part of seismic wave motion. Rotations can be important in earthquake strong motion and in Induced Seismicity Monitoring. Rotational seismic data can also enable shear selectivity and improve wavefield sampling for vertical geophones in 3D surveys, among other applications. However, sensor technology has been a limiting factor to date. The US Department of Energy (DOE) and Applied Technology Associates (ATA) are funding a multi-year project that is now entering Phase 2 to develop and deploy a new generation of rotational sensors for validation of rotational seismic applications. Initial focus is on induced seismicity monitoring, particularly for Enhanced Geothermal Systems (EGS) with fracturing. The sensors employ Magnetohydrodynamic (MHD) principles with broadband response, improved noise floors, robustness, and repeatability. This paper presents a summary of Phase 1 results and Phase 2 status.

  8. Flow past a rotating cylinder

    Science.gov (United States)

    Mittal, Sanjay; Kumar, Bhaskar

    2003-02-01

    Flow past a spinning circular cylinder placed in a uniform stream is investigated via two-dimensional computations. A stabilized finite element method is utilized to solve the incompressible Navier Stokes equations in the primitive variables formulation. The Reynolds number based on the cylinder diameter and free-stream speed of the flow is 200. The non-dimensional rotation rate, [alpha] (ratio of the surface speed and freestream speed), is varied between 0 and 5. The time integration of the flow equations is carried out for very large dimensionless time. Vortex shedding is observed for [alpha] cylinder. The results from the stability analysis for the rotating cylinder are in very good agreement with those from direct numerical simulations. For large rotation rates, very large lift coefficients can be obtained via the Magnus effect. However, the power requirement for rotating the cylinder increases rapidly with rotation rate.

  9. Bidirectional optical rotation of cells

    Directory of Open Access Journals (Sweden)

    Jiyi Wu

    2017-08-01

    Full Text Available Precise and controlled rotation manipulation of cells is extremely important in biological applications and biomedical studies. Particularly, bidirectional rotation manipulation of a single or multiple cells is a challenge for cell tomography and analysis. In this paper, we report an optical method that is capable of bidirectional rotation manipulation of a single or multiple cells. By launching a laser beam at 980 nm into dual-beam tapered fibers, a single or multiple cells in solutions can be trapped and rotated bidirectionally under the action of optical forces. Moreover, the rotational behavior can be controlled by altering the relative distance between the two fibers and the input optical power. Experimental results were interpreted by numerical simulations.

  10. Novel configuration for an enhanced and compact all-fiber Faraday rotator with matched birefringence.

    Science.gov (United States)

    Asraf, Sagie; Sintov, Yoav; Zalevsky, Zeev

    2017-08-07

    We propose a novel configuration for an improved and compact all fiber Faraday rotator based on phase matching between the Faraday rotation and bend-induced birefringence. The device utilizes a coiled fiber within two electro-magnetic toroids, such that the fiber length required for getting the beat length is quite long and several rounds of fiber are needed. Analysis of the capabilities of the proposed device and its sensitivity to different parameters is presented. Faraday rotation of 13° was experimentally measured in six meters of single mode silica fiber, with a magnetic field of about 0.06T at a wavelength of 1064nm. We show that phase matching between the two phenomena significantly improves the polarization rotation by a factor of 4-10. In addition, we demonstrate the ability to achieve higher rotation by using Fabry Perot resonator in low terbium doped glass.

  11. Faraday rotation due to excitation of magnetoplasmons in graphene microribbons.

    Science.gov (United States)

    Tymchenko, Mykhailo; Nikitin, Alexey Yu; Martín-Moreno, Luis

    2013-11-26

    A single graphene sheet, when subjected to a perpendicular static magnetic field, provides a Faraday rotation that, per atomic layer, greatly surpasses that of any other known material. In continuous graphene, Faraday rotation originates from the cyclotron resonance of massless carriers, which allows dynamical tuning through either external electrostatic or magneto-static setting. Furthermore, the rotation direction can be controlled by changing the sign of the carriers in graphene, which can be done by means of an external electric field. However, despite these tuning possibilities, the requirement of large magnetic fields hinders the application of the Faraday effect in real devices, especially for frequencies higher than a few terahertz. In this work we demonstrate that large Faraday rotation can be achieved in arrays of graphene microribbons, through the excitation of the magnetoplasmons of individual ribbons, at larger frequencies than those dictated by the cyclotron resonance. In this way, for a given magnetic field and chemical potential, structuring graphene periodically can produce large Faraday rotation at larger frequencies than what would occur in a continuous graphene sheet. Alternatively, at a given frequency, graphene ribbons produce large Faraday rotation at much smaller magnetic fields than in continuous graphene.

  12. Quality assurance of MLC leaf position accuracy and relative dose effect at the MLC abutment region using an electronic portal imaging device

    International Nuclear Information System (INIS)

    Sumida, Iori; Yamaguchi, Hajime; Kizaki, Hisao; Koizumi, Masahiko; Ogata, Toshiyuki; Takahashi, Yutaka; Yoshioka, Yasuo

    2012-01-01

    We investigated an electronic portal image device (EPID)-based method to see whether it provides effective and accurate relative dose measurement at abutment leaves in terms of positional errors of the multi-leaf collimator (MLC) leaf position. A Siemens ONCOR machine was used. For the garden fence test, a rectangular field (0.2x20 cm) was sequentially irradiated 11 times at 2-cm intervals. Deviations from planned leaf positions were calculated. For the nongap test, relative doses at the MLC abutment region were evaluated by sequential irradiation of a rectangular field (2x20 cm) 10 times with a MLC separation of 2 cm without a leaf gap. The integral signal in a region of interest was set to position A (between leaves) and B (neighbor of A). A pixel value at position B was used as background and the pixel ratio (A/Bx100) was calculated. Both tests were performed at four gantry angles (0, 90, 180 and 270deg) four times over 1 month. For the nongap test the difference in pixel ratio between the first and last period was calculated. Regarding results, average deviations from planned positions with the garden fence test were within 0.5 mm at all gantry angles, and at gantry angles of 90 and 270deg tended to decrease gradually over the month. For the nongap test, pixel ratio tended to increase gradually in all leaves, leading to a decrease in relative doses at abutment regions. This phenomenon was affected by both gravity arising from the gantry angle, and the hardware-associated contraction of field size with this type of machine. (author)

  13. Hydrodynamics of rotating superfluids

    International Nuclear Information System (INIS)

    Chandler, E.A.

    1981-01-01

    In this thesis, a coarse grained hydrodynamics is developed from the exact description of Tkachenko. To account for the dynamics of the vortex lattice, the macroscopic vortex displacement field is treated as an independent degree of freedom. The conserved energy is written in terms of the coarse-grained normal fluid, superfluid, and vortex velocities and includes an elastic energy associated with deformations of the vortex lattice. Equations of motion consistent with the conservation of energy, entropy and vorticity and containing mutual friction terms arising from microscopic interactions between normal fluid excitations and the vortex lines are derived. When the vortex velocity is eliminated from the damping terms, this system of equations becomes essentially that of BK with added elastic terms in the momentum stress tensor and energy current. The dispersion relation and damping of the first and second sound modes and the two transverse modes sustained by the system are investigated. It is shown that mutual friction mixes the transverse modes of the normal and superfluid components and damps the transverse mode associated with the relative velocity of these components, making this wave evanescent in the plane perpendicular to the rotation axis. The wave associated with transverse motion of the total mass current is a generalized Tkachenko mode, whose dispersion relation reduces to that derived by Tkachenko wave when the wavevector lies in this plane

  14. Magnetostrophic Rotating Magnetoconvection

    Science.gov (United States)

    King, Eric; Aurnou, Jonathan

    2016-11-01

    Planetary magnetic fields are generated by turbulent convection within their vast interior liquid metal cores. Although direct observation is not possible, this liquid metal circulation is thought to be dominated by the controlling influences of Coriolis and Lorentz forces. Theory famously predicts that local-scale convection naturally settles into the so-called magnetostrophic state, where the Coriolis and Lorentz forces partially cancel, and convection is optimally efficient. To date, no laboratory experiments have reached the magnetostrophic regime in turbulent liquid metal convection. Furthermore, computational dynamo simulations have as yet failed to produce a globally magnetostrophic dynamo, which has led some to question the existence of the magnetostrophic state. Here, we present results from the first turbulent magnetostrophic rotating magnetoconvection experiments using the liquid metal gallium. We find that turbulent convection in the magnetostrophic regime is, in fact, maximally efficient. The experimental results clarify these previously disparate results, suggesting that the fluid dynamics saturate in magnetostrophic balance within turbulent liquid metal, planetary cores. The authors thank the NSF Geophysics Program for financial support.

  15. Rotating quantum Gaussian packets

    International Nuclear Information System (INIS)

    Dodonov, V V

    2015-01-01

    We study two-dimensional quantum Gaussian packets with a fixed value of mean angular momentum. This value is the sum of two independent parts: the ‘external’ momentum related to the motion of the packet center and the ‘internal’ momentum due to quantum fluctuations. The packets minimizing the mean energy of an isotropic oscillator with the fixed mean angular momentum are found. They exist for ‘co-rotating’ external and internal motions, and they have nonzero correlation coefficients between coordinates and momenta, together with some (moderate) amount of quadrature squeezing. Variances of angular momentum and energy are calculated, too. Differences in the behavior of ‘co-rotating’ and ‘anti-rotating’ packets are shown. The time evolution of rotating Gaussian packets is analyzed, including the cases of a charge in a homogeneous magnetic field and a free particle. In the latter case, the effect of initial shrinking of packets with big enough coordinate-momentum correlation coefficients (followed by the well known expansion) is discovered. This happens due to a competition of ‘focusing’ and ‘de-focusing’ in the orthogonal directions. (paper)

  16. Rotator cuff pathology

    International Nuclear Information System (INIS)

    Pigeau, I.; Doursounian, L.; Maigne, J.Y.; Guinet, C.; Meary, E.; Buy, J.N.; Touzard, R.C.; Vadrot, D.; Laval-Jeantet, M.

    1989-01-01

    Fifteen volunteers and 73 patients with suspected rotator cuff lesions were examined at 0.5 T with T2 * -weighted gradient-echo (GE) MR imaging (700/33/30 degrees) (oblique coronal and sagittal 3 mm thick, surface coil). Results were compared with those of arthrography (all cases), T1-weighted GE imaging (400/20/90 degrees) (35 cases), surgery (28 cases), and T2-weighted spin-echo (SE) images (2,000/60-120) (17 cases). GE images demonstrated all tears (complete, 32, partial, 12) and was superior to arthrography in determining site and size and in displaying muscles (critical point in surgical planning). In 20 cases without tears on arthrography, GE imaging demonstrated five cases of tendinitis, five cases of bursitis, and six probable intratendinous or superficial partial tears. T2 * -weighted GE imaging was superior to T2-weighted SE and T1-weighted GE imaging, with higher fluid contrast and a low fat signal. Therefore, it might replace arthrography in the diagnosis and surgical approach to this pathology

  17. System and Method for Obtaining Simultaneous Levitation and Rotation of a Ferromagnetic Object

    Science.gov (United States)

    Banerjee, Subrata; Sarkar, Mrinal Kanti; Ghosh, Arnab

    2017-02-01

    In this work a practical demonstration for simultaneous levitation and rotation for a ferromagnetic cylindrical object is presented. A hollow steel cylinder has been arranged to remain suspended stably under I-core electromagnet utilizing dc attraction type levitation principle and then arranged to rotate the levitated object around 1000 rpm speed based on eddy current based energy meter principle. Since the object is to be rotating during levitated condition the device will be frictionless, energy-efficient and robust. This technology may be applied to frictionless energy meter, wind turbine, machine tool applications, precision instruments and many other devices where easy energy-efficient stable rotation will be required. The cascade lead compensation control scheme has been applied for stabilization of unstable levitation system. The proposed device is successfully tested in the laboratory and experimental results have been produced.

  18. Thermonuclear device

    International Nuclear Information System (INIS)

    Tezuka, Masaru.

    1993-01-01

    Protrusions and recesses are formed to a vacuum vessel and toroidal magnetic coils, and they are engaged. Since the vacuum vessel is generally supported firmly by a rack or the like by support legs, the toroidal magnetic field coils can be certainly supported against tumbling force. Then, there can be attained strong supports for the toroidal magnetic field coils, in addition to support by wedges on the side of inboard and support by share panels on the side of outboard, capable of withstanding great electromagnetic forces which may occur in large-scaled next-generation devices. That is, toroidal magnetic field coils excellent from a view point of deformation and stress can be obtained, to provide a thermonuclear device of higher reliability. (N.H.)

  19. Thermonuclear device

    International Nuclear Information System (INIS)

    Oosaki, Osamu; Masuda, Kenju.

    1980-01-01

    Purpose: To provide excellent electric properties and high reliability in a thermonuclear device by improving a current collecting board connected to a coil device. Constitution: A current collecting board element perforated with an opening for enserting a connecting terminal is sized to be inserted into a plating tank, and is surface treated in the plating tank. Only the current collecting board element preferably surface treated is picked up. A plurality of such current collecting board elements are connected and welded to form a large current collecting board. In this manner, the current collecting board having several m 2 to several ten order m 2 in area can be obtained as preferably surface treated at the connecting terminal hole. The current collecting board element can be determined in shape with the existing facility without increasing the size of a surface treating tank. (Kamimura, M.)

  20. Thermonuclear device

    International Nuclear Information System (INIS)

    Yagi, Yasuomi; Takahashi, Ken; Hashimoto, Hiroshi.

    1984-01-01

    Purpose: To improve the plasma confining performances by bringing the irregular magnetic fields nearly to zero and decreasing the absolute value of the irregular magnetic fields at every positions. Constitution: The winding direction of a plurality of coil elements, for instance, double pan cake coils of toroidal coils in a torus type or mirror type thermonuclear device are reversed to each other in their laminating direction, whereby the irregular magnetic fields due to the coil-stepped portions in each toroidal coils are brought nearly to zero. This enables to bring the average irregular magnetic fields as a whole in the thermonuclear device nearly to zero, as well as, decrease the absolute value of the irregular magnetic fields in each positions. Thus, the plasma confining performances can be improved. (Moriyama, K.)

  1. Ultrasonic viewing device

    International Nuclear Information System (INIS)

    Ito, Juro.

    1979-01-01

    Purpose: To improve the safety of reactor operation by enabling to detect the states and positions of fuel assemblies over a wide range with a set of ultrasonic viewing device comprising a rotatable ultrasonic transmitter-receiver and a reflector mounted with an adjustable angle. Constitution: A driving portion for a ultrasonic viewing device is provided to a rotary plug closing the opening of a reactor vessel and a guide pipe suspending below the coolant level is provided to the driving portion. An ultrasonic transmitter-receiver is provided at the end of the holder tube in the guide pipe. A reflector is provided at the upper position of the reactor core so as to correspond to the ultrasonic transmitter-receiver. The ultrasonic transmitter-receiver, positioned by the driving portion, performs horizontal movement for scanning the entire surface of the top of the reactor core, as well as vertical movement covering the gap between the upper mechanism on the reactor and the reactor core, whereby the confirmation for the separation of the control rod and the detection for the states of the reactor core can be conducted by the reflection waves from the reflector. (Moriyama, K.)

  2. Determination of the optimal speed of rotational display through an 180 degree arc in rotatostereoradiography and MR angiography

    International Nuclear Information System (INIS)

    Ottomo, M.; Takekawa, S.D.; Sugawara, K.; Nakamura, T.; Fujimoto, M.; Nakanishi, T.

    1990-01-01

    Rotatostereoradiographic (RSRG) images are displayed in an oscillating, rotational manner. While reviewing these rotating images, the radiologist may become psychologically irritated by the rotation. A rapidly rotating display of linear subjects gives one three-dimensional depth information. This three-dimensional sense is lost if the rotation speed is too slow. The authors of this paper determined the slowest possible rotating display speed that allows perception of three-dimensional depth information minimizing psychological irritation. In the RSRG device (Shimadzu ROTATO-360), an x-ray tube coupled with an image intensifier rotates through a 180 degrees arc in 1.8 or 2.25 seconds. Both rotation times could be doubled. The images were displayed at four different speeds, covering the 180 degrees arc in 1.8, 2.25, 3.6, and 4.5 seconds

  3. "Distinvar" device

    CERN Multimedia

    CERN PhotoLab

    1965-01-01

    The alignment of one of the accelerator magnets being checked by the AR Division survey group. A "distinvar" device, invented by the group, using calibrated invar wires stretched between the fixed survey pillar (on the left) and a fixed point on the magnet. In two days it is thus possible to measure the alignment of the 100 magnets with an accuracy better than 1/10.

  4. Latching device

    Science.gov (United States)

    Ulrich, G. W. (Inventor)

    1975-01-01

    A latching device is suited for use in establishing a substantially motionless connection between a stationary receiver and a movable latching mechanism. The latching mechanism includes a pivotally supported restraining hook continuously urged into a capturing relationship with the receiver, characterized by a spring-biased pawl having a plurality of aligned teeth. The teeth are seated in the surface of the throat of the hook and positionable into restraining engagement with a rigid restraining shoulder projected from the receiver.

  5. Rotational discontinuities in anisotropic plasmas

    International Nuclear Information System (INIS)

    Omidi, N.

    1992-01-01

    The kinetic structure of rotational discontinuities (RDs) in anisotropic plasmas with T perpendicular /T parallel > 1 is investigated by using a one-dimensional electromagnetic hybrid code. To form the RD, a new approach is used where the plasma is injected from one boundary and reflected from the other, resulting in the generation of a traveling fast shock and an RD. Unlike the previously used methods, no a priori assumptions are made regarding the initial structure (i.e. width or sense of rotation) of the rotational discontinuity. The results show that across the RD both the magnetic field strength and direction, as well as the plasma density change. Given that such a change can also be associated with an intermediate shock, the Rankine-Hugoniot relations are used to confirm that the observed structures are indeed RDs. It is found that the thickness of RDs is a few ion inertial lengths and is independent of the rotation angle. Also, the preferred sense of rotation is in the electron sense; however, RDs with a rotation angle larger than 180 degree are found to be unstable, changing their rotation to a stable ion sense

  6. [Rotator cuff tear athropathy prevalence].

    Science.gov (United States)

    Guerra-Soriano, F; Encalada-Díaz, M I; Ruiz-Suárez, M; Valero-González, F S

    2017-01-01

    Glenohumeral arthritis secondary to massive rotator cuff tear presents with a superior displacement and femoralization of the humeral head with coracoacromial arch acetabularization. The purpose of this study was to establish prevalence of rotator cuff tear artropathy (CTA) at our institution. Four hundred electronic records were reviewed from which we identified 136 patients with rotator cuff tears. A second group was composed with patients with massive cuff tears that were analized and staged by the Seebauer cuff tear arthropathy classification. Thirty four patients with massive rotator cuff tears were identified, 8 male and 26 female (age 60.1 ± 10.26 years). Massive rotator cuff tear prevalence was 25%. CTA prevalence found in the rotator cuff group was 19 and 76% in the massive cuff tears group. Patients were staged according to the classification with 32% in stage 1a, 11% 1b, 32% 2a and 0% 2b. CTA prevalence in patients with rotator cuff tears and massive cuff tears is higher than the one reported in American population. We consider that a revision of the Seebauer classification to be appropriate to determine its reliability.

  7. Rotating relativistic neutron stars

    Energy Technology Data Exchange (ETDEWEB)

    Weber, F.; Glendenning, N.K.

    1991-07-21

    Models of rotating neutron stars are constructed in the framework of Einstein's theory of general relativity. For this purpose a refined version of Hartle's method is applied. The properties of these objects, e.g. gravitational mass, equatorial and polar radius, eccentricity, red- and blueshift, quadrupole moment, are investigated for Kepler frequencies of 4000 s{sup {minus}1} {le} {Omega}{sub K} {le} 9000 s{sup {minus}1}. Therefore a self-consistency problem inherent in the determination of {Omega}{sub K} must be solved. The investigation is based on neutron star matter equations of state derived from the relativistic Martin-Schwinger hierarch of coupled Green's functions. By means of introducing the Hartree, Hartree-Fock, and ladder ({Lambda}) approximations, models of the equation of state derived. A special feature of the latter approximation scheme is the inclusion of dynamical two-particle correlations. These have been calculated from the relativistic T-matrix applying both the HEA and Bonn meson-exchange potentials of the nucleon-nucleon force. The nuclear forces of the former two treatments are those of the standard scalar-vector-isovector model of quantum hadron dynamics, with parameters adjusted to the nuclear matter data. An important aspect of this work consists in testing the compatibility of different competing models of the nuclear equation of state with data on pulsar periods. By this the fundamental problem of nuclear physics concerning the behavior of the equation of state at supernuclear densities can be treated.

  8. Fuel exchange device for FBR type reactor

    International Nuclear Information System (INIS)

    Onuki, Koji.

    1993-01-01

    The device of the present invention can provide fresh fuels with a rotational angle aligned with the direction in the reactor core, so that the fresh fuels can be inserted being aligned with apertures of the reactor core even if a self orientation mechanism should fail to operate. That is, a rotational angle detection means (1) detects the rotational angle of fresh fuels before insertion to the reactor core. A fuel rotational angle control means (2) controls the rotational angle of the fresh fuels by comparing the detection result of the means (1) and the data for the insertion position of the reactor core. A fuel rotation means (3) compensates the rotational angel of the fresh fuels based on the control signal from the means (2). In this way, when the fresh fuels are inserted to the reactor core, the fresh fuels set at the same angle as that for the aperture of the reactor core. Accordingly, even if the self orientation mechanism should not operate, the fresh fuels can be inserted smoothly. As a result, it is possible to save loss time upon fuel exchange and mitigate operator's burden during operation. (I.S.)

  9. Soft, Rotating Pneumatic Actuator.

    Science.gov (United States)

    Ainla, Alar; Verma, Mohit S; Yang, Dian; Whitesides, George M

    2017-09-01

    This article describes a soft pneumatic actuator that generates cyclical motion. The actuator consists of several (three, four, or five) chambers (arranged around the circumference of a circle surrounding a central rod) that can be actuated independently using negative pressure (or partial vacuum). Sequential actuation of the four-chamber device using reduced pressure moves the central rod cyclically in an approximately square path. We characterize the trajectory of the actuator and the force exerted by it, as we vary the material used for fabrication, the number of chambers, and the size of the actuator. We demonstrate two applications of this actuator: to deliver fluid while stirring (by replacing the central rod with a needle) and for locomotion that mimics a reptilian gait (by combining four actuators together).

  10. Effect of rotation on convective mass transfer in rotating channels

    International Nuclear Information System (INIS)

    Pharoah, J.G.; Djilali, N.

    2002-01-01

    Laminar flow and mass transfer in rotating channels is investigated in the context of centrifugal membrane separation. The effect of orientation with respect to the rotational axis is examined for rectangular channels of aspect ratio 3 and the Rossby number is varied from 0.3 to 20.9. Both Ro and the channel orientation are found to have a significant effect on the flow. Mass transfer calculations corresponding to reverse osmosis desalination are carried out at various operating pressures and all rotating cases exhibit significant process enhancements at relatively low rotation rates. Finally, while it is common in the membrane literature to correlate mass transfer performance with membrane shear rates this is shown not to be valid in the cases presented herein. (author)

  11. Dosimetric Changes Resulting From Patient Rotational Setup Errors in Proton Therapy Prostate Plans

    International Nuclear Information System (INIS)

    Sejpal, Samir V.; Amos, Richard A.; Bluett, Jaques B.; Levy, Lawrence B.; Kudchadker, Rajat J.; Johnson, Jennifer; Choi, Seungtaek; Lee, Andrew K.

    2009-01-01

    Purpose: To evaluate the dose changes to the target and critical structures from rotational setup errors in prostate cancer patients treated with proton therapy. Methods and Materials: A total of 70 plans were analyzed for 10 patients treated with parallel-opposed proton beams to a dose of 7,600 60 Co-cGy-equivalent (CcGE) in 200 CcGE fractions to the clinical target volume (i.e., prostate and proximal seminal vesicles). Rotational setup errors of +3 o , -3 deg., +5 deg., and -5 deg. (to simulate pelvic tilt) were generated by adjusting the gantry. Horizontal couch shifts of +3 deg. and -3 deg. (to simulate longitudinal setup variability) were also generated. Verification plans were recomputed, keeping the same treatment parameters as the control. Results: All changes shown are for 38 fractions. The mean clinical target volume dose was 7,780 CcGE. The mean change in the clinical target volume dose in the worse case scenario for all shifts was 2 CcGE (absolute range in worst case scenario, 7,729-7,848 CcGE). The mean changes in the critical organ dose in the worst case scenario was 6 CcGE (bladder), 18 CcGE (rectum), 36 CcGE (anterior rectal wall), and 141 CcGE (femoral heads) for all plans. In general, the percentage of change in the worse case scenario for all shifts to the critical structures was <5%. Deviations in the absolute percentage of volume of organ receiving 45 and 70 Gy for the bladder and rectum were <2% for all plans. Conclusion: Patient rotational movements of 3 deg. and 5 deg. and horizontal couch shifts of 3 deg. in prostate proton planning did not confer clinically significant dose changes to the target volumes or critical structures.

  12. Spectroscopic studies of a high Mach-number rotating plasma flow

    International Nuclear Information System (INIS)

    Ando, Akira; Ashino, Masashi; Sagi, Yukiko; Inutake, Masaaki; Hattori, Kunihiko; Yoshinuma, Mikirou; Imasaki, Atsushi; Tobari, Hiroyuki; Yagai, Tsuyoshi

    2001-01-01

    Characteristics of an axially-magnetized rotating plasma are investigated by spectroscopy in the HITOP device of Tohoku University. A He plasma flows our axially and rotates azimuthally near the muzzle region of the MPD arcjet. Flow and rotational velocities and temperature of He ions and atoms are measured by Doppler shift and broadening of the HeII (γ=468.58 nm) and HeI (γ=587.56 nm) lines. Rotational velocity increases with the increase of axially-applied magnetic field strength and discharge current. As discharge current increases and mass flow rate decreases, the plasma flow velocity increases and T i increases. Ion acoustic Mach number of the plasma flow also increases, but tends to saturate at near 1. Radial profile of space potential is calculated from the obtained rotational velocity. The potential profile in the core region is parabolic corresponding to the observed rigid-body rotation of the core plasma. (author)

  13. Spectroscopic studies of a high Mach-number rotating plasma flow

    Energy Technology Data Exchange (ETDEWEB)

    Ando, Akira; Ashino, Masashi; Sagi, Yukiko; Inutake, Masaaki; Hattori, Kunihiko; Yoshinuma, Mikirou; Imasaki, Atsushi; Tobari, Hiroyuki; Yagai, Tsuyoshi [Tohoku Univ., Dept. of Electrical Engineering, Sendai, Miyagi (Japan)

    2001-07-01

    Characteristics of an axially-magnetized rotating plasma are investigated by spectroscopy in the HITOP device of Tohoku University. A He plasma flows our axially and rotates azimuthally near the muzzle region of the MPD arcjet. Flow and rotational velocities and temperature of He ions and atoms are measured by Doppler shift and broadening of the HeII ({gamma}=468.58 nm) and HeI ({gamma}=587.56 nm) lines. Rotational velocity increases with the increase of axially-applied magnetic field strength and discharge current. As discharge current increases and mass flow rate decreases, the plasma flow velocity increases and T{sub i} increases. Ion acoustic Mach number of the plasma flow also increases, but tends to saturate at near 1. Radial profile of space potential is calculated from the obtained rotational velocity. The potential profile in the core region is parabolic corresponding to the observed rigid-body rotation of the core plasma. (author)

  14. Axial gap rotating electrical machine

    Science.gov (United States)

    None

    2016-02-23

    Direct drive rotating electrical machines with axial air gaps are disclosed. In these machines, a rotor ring and stator ring define an axial air gap between them. Sets of gap-maintaining rolling supports bear between the rotor ring and the stator ring at their peripheries to maintain the axial air gap. Also disclosed are wind turbines using these generators, and structures and methods for mounting direct drive rotating electrical generators to the hubs of wind turbines. In particular, the rotor ring of the generator may be carried directly by the hub of a wind turbine to rotate relative to a shaft without being mounted directly to the shaft.

  15. On the relativity of rotation

    International Nuclear Information System (INIS)

    Gron, O.

    2010-01-01

    The question whether rotational motion is relative according to the general theory of relativity is discussed. Einstein's ambivalence concerning this question is pointed out. In the present article I defend Einstein's way of thinking on this when he presented the theory in 1916. The significance of the phenomenon of perfect inertial dragging in connection with the relativity of rotational motion is discussed. The necessity of introducing an extended model of the Minkowski spacetime, in which a globally empty space is supplied with a cosmic mass shell with radius equal to its own Schwarzschild radius, in order to extend the principle of relativity to accelerated and rotational motion, is made clear.

  16. Device and method for redirecting electromagnetic signals

    Science.gov (United States)

    Garcia, Ernest J.

    1999-01-01

    A device fabricated to redirect electromagnetic signals, the device including a primary driver adapted to provide a predetermined force, a linkage system coupled to the primary driver, a pusher rod rotationally coupled to the linkage system, a flexible rod element attached to the pusher rod and adapted to buckle upon the application of the predetermined force, and a mirror structure attached to the flexible rod element at one end and to the substrate at another end. When the predetermined force buckles the flexible rod element, the mirror structure and the flexible rod element both move to thereby allow a remotely-located electromagnetic signal directed towards the device to be redirected.

  17. Regenerative-filter-incinerator device

    Energy Technology Data Exchange (ETDEWEB)

    Rosebrock, T.L.

    1977-10-18

    A regenerative-filter-incinerator device, for use in the exhaust system of a diesel engine, includes a drum-like regenerative-heat exchanger-filter assembly rotatably mounted within a housing that is adapted to be installed directly in the exhaust gas stream discharged from a diesel engine as close to the engine as possible. The regenerative-heat exchanger-filter assembly provides an inner chamber which serves as a reaction chamber for the secondary combustion of exhaust gases including particulates discharged from the engine. The regenerative-heat exchanger-filter assembly includes separately rotatable heat exchange-filter elements pervious to radial flow of fluid therethrough and adapted to filter out particulates from the exhaust gases and to carry them into the reaction chamber. During engine operation, the reaction chamber is provided with a quantity of heat, as necessary, to effect secondary combustion of the exhaust gases and particulates by means of an auxiliary heat source and the heat generated within the reaction chamber is stored in the individual heat exchange-filter elements during the discharge of exhaust gases therethrough from the reaction chamber and this heat is then transferred to the inflowing volume of the exhaust gases so that, in effect, exhaust gas is discharged from the device at substantially the same temperature as it was during its inlet into the device from the engine.

  18. Rotational microfluidic motor for on-chip microcentrifugation

    Science.gov (United States)

    Shilton, Richie J.; Glass, Nick R.; Chan, Peggy; Yeo, Leslie Y.; Friend, James R.

    2011-06-01

    We report on the design of a surface acoustic wave (SAW) driven fluid-coupled micromotor which runs at high rotational velocities. A pair of opposing SAWs generated on a lithium niobate substrate are each obliquely passed into either side of a fluid drop to drive rotation of the fluid, and the thin circular disk set on the drop. Using water for the drop, a 5 mm diameter disk was driven with rotation speeds and start-up torques up to 2250 rpm and 60 nN m, respectively. Most importantly for lab-on-a-chip applications, radial accelerations of 172 m/s2 was obtained, presenting possibilities for microcentrifugation, flow sequencing, assays, and cell culturing in truly microscale lab-on-a-chip devices.

  19. Sensorless Estimation and Nonlinear Control of a Rotational Energy Harvester

    Science.gov (United States)

    Nunna, Kameswarie; Toh, Tzern T.; Mitcheson, Paul D.; Astolfi, Alessandro

    2013-12-01

    It is important to perform sensorless monitoring of parameters in energy harvesting devices in order to determine the operating states of the system. However, physical measurements of these parameters is often a challenging task due to the unavailability of access points. This paper presents, as an example application, the design of a nonlinear observer and a nonlinear feedback controller for a rotational energy harvester. A dynamic model of a rotational energy harvester with its power electronic interface is derived and validated. This model is then used to design a nonlinear observer and a nonlinear feedback controller which yield a sensorless closed-loop system. The observer estimates the mechancial quantities from the measured electrical quantities while the control law sustains power generation across a range of source rotation speeds. The proposed scheme is assessed through simulations and experiments.

  20. Guidewire-Controlled Advancement of the Amplatz Thrombectomy Device

    International Nuclear Information System (INIS)

    Mueller-Huelsbeck, Stefan; Schwarzenberg, Helmut; Heller, Martin

    1998-01-01

    The Amplatz Thrombectomy Device (ATD) is a percutaneous rotational catheter proven to homogenize thrombus. The catheter design allows neither application over a coaxial running guidewire nor the use of the device as a monorail system. We report a technical modification that provides guided advancement of the catheter over a wire in order to prevent failure of application and to facilitate the interventional procedure

  1. Device for verification of seals by optical fibers

    International Nuclear Information System (INIS)

    Neuilly, M.; Guitaud, M.H.; Dufour, J.L.

    1993-01-01

    The seal is built around an optical fiber with a verification device. This device is made of a lighting source and photographic means supported by a rotating assembly which allow to take an image of each optical fiber end. These images are compared with reference images taken just after the sealing. (A.B.). 5 refs., 6 figs

  2. Spent fuel container alignment device and method

    Science.gov (United States)

    Jones, Stewart D.; Chapek, George V.

    1996-01-01

    An alignment device is used with a spent fuel shipping container including a plurality of fuel pockets for spent fuel arranged in an annular array and having a rotatable cover including an access opening therein. The alignment device includes a lightweight plate which is installed over the access opening of the cover. A laser device is mounted on the plate so as to emit a laser beam through a laser admittance window in the cover into the container in the direction of a pre-established target associated with a particular fuel pocket. An indexing arrangement on the container provides an indication of the angular position of the rotatable cover when the laser beam produced by the laser is brought into alignment with the target of the associated fuel pocket.

  3. On rapid rotation in stellarators

    International Nuclear Information System (INIS)

    Helander, Per

    2008-01-01

    The conditions under which rapid plasma rotation may occur in a three-dimensional magnetic field, such as that of a stellarator, are investigated. Rotation velocities comparable to the ion thermal speed are found to be attainable only in magnetic fields which are approximately isometric. In an isometric magnetic field the dependence of the magnetic field strength B on the arc length l along the field is the same for all field lines on each flux surface ψ. Only in fields where the departure from exact isometry, B=B(ψ,l), is of the order of the ion gyroradius divided by the macroscopic length scale are rotation speeds comparable to the ion thermal speed possible. Moreover, it is shown that the rotation must be in the direction of the vector ∇ψx∇B. (author)

  4. Spontaneous Rotational Inversion in Phycomyces

    KAUST Repository

    Goriely, Alain

    2011-03-01

    The filamentary fungus Phycomyces blakesleeanus undergoes a series of remarkable transitions during aerial growth. During what is known as the stagea IV growth phase, the fungus extends while rotating in a counterclockwise manner when viewed from above (stagea IVa) and then, while continuing to grow, spontaneously reverses to a clockwise rotation (stagea IVb). This phase lasts for 24-48Ah and is sometimes followed by yet another reversal (stageAIVc) before the overall growth ends. Here, we propose a continuum mechanical model of this entire process using nonlinear, anisotropic, elasticity and show how helical anisotropy associated with the cell wall structure can induce spontaneous rotation and, under appropriate circumstances, the observed reversal of rotational handedness. © 2011 American Physical Society.

  5. Differential rotation in magnetic stars

    International Nuclear Information System (INIS)

    Moss, D.

    1981-01-01

    The possibility that large-scale magnetic fields in stars are the product of a contemporary dynamo situated in the convective stellar core, rather than being a fossil from an earlier stage in the history of the star, is investigated. It is demonstrated that then the envelope will almost inevitably be in a state of differential rotation. Some simple models are constructed to illustrate the magnitude of the effects on the structure of the envelope and magnetic field. It is found that, for models which are relatively rapidly rotating, a modest differential rotation at the surface of the core may increase considerably the ratio of internal to surface field, but only give rise to a small surface differential rotation. (author)

  6. Conjunct rotation: Codman's paradox revisited.

    Science.gov (United States)

    Wolf, Sebastian I; Fradet, Laetitia; Rettig, Oliver

    2009-05-01

    This contribution mathematically formalizes Codman's idea of conjunct rotation, a term he used in 1934 to describe a paradoxical phenomenon arising from a closed-loop arm movement. Real (axial) rotation is distinguished from conjunct rotation. For characterizing the latter, the idea of reference vector fields is developed to define the neutral axial position of the humerus for any given orientation of its long axis. This concept largely avoids typical coordinate singularities arising from decomposition of 3D joint motion and therefore can be used for postural (axial) assessment of the shoulder joint both clinically and in sports science in almost the complete accessible range of motion. The concept, even though algebraic rather complex, might help to get an easier and more intuitive understanding of axial rotation of the shoulder in complex movements present in daily life and in sports.

  7. Data on rotating plasma operation in Tornado traps

    International Nuclear Information System (INIS)

    Lehnert, B.

    1976-01-01

    Rotating plasma operation in spiral coil 'Tornado' traps provides a unique combination of confinement and heating properties. Such a system consists of a closed and compact magnetic bottle to which the crossed-field technique can be applied, in absence of end insulators and their critical velocity limitation effect. This should make possible the generation and heating of fully ionized plasmas by simple means, within a large range of ion densities, temperatures, and rotational velocities. Provided that stable operation becomes possible at high temperatures, it is likely that Tornado traps can be used as strong neutron sources of moderately large dimensions and technically realizable parameter values. Some detailed data and operation ranges are given for the 'Tornado 650' device in Leningrad, and for a somewhat larger device to be operated as neutron source. (Auth.)

  8. Data on rotating plasma operation in Tornado traps

    International Nuclear Information System (INIS)

    Lehnert, B.

    1977-01-01

    Rotating plasma operation in spiral coil 'Tornado' traps provides a unique combination of confinement and heating properties. Such a system consists of a closed and compact magnetic bottle to which the crossed-field technique can be applied, in absence of end insulators and their critical velocity limitation effect. This is expected to lead to the generation and heating of fully ionized plasmas within a large range of ion densities, temperatures, and rotational velocities. Provided that stable operation becomes possible at high temperatures and the effects due to the asymmetries of the spiral coil structure can be neglected, it is likely that Tornado traps can be used as strong neutron sources of moderately large dimensions and technically realizable parameter values. Some detailed data and operation ranges are outlined for the 'Tornado 650' device in Leningrad, and for a somewhat larger device which may be operated as a neutron source. (Auth.)

  9. A Tool for Simulating Rotating Coil Magnetometers

    CERN Document Server

    Bottura, L; Schnizer, P; Smirnov, N

    2002-01-01

    When investigating the quality of a magnetic measurement system, one observes difficulties to identify the "trouble maker" of such a system as different effects can yield similar influences on the measurement results.We describe a tool in this paper that allows to investigate numerically the effects produced by different imperfections of components of such a system, including, but not limited to vibration and movements of the rotating coil, influence of electrical noise on the system, angular encoder imperfections. This system can simulate the deterministic and stochastic parts of those imperfections. We outline the physical models used that are generally based on experience or first principles. Comparisons to analytical results are shown. The modular structure of the general design of this tool permits to include new modules for new devices and effects.

  10. Damping device

    International Nuclear Information System (INIS)

    Banks, E.L. Jr.; Dowell, T.P.

    1976-01-01

    The description is given of a damper which includes a pair of telescopic components interconnected by relative linear movement one in relation to the other, by a screw and ball nut device, with a braking surface on one of the components, a brake engaging the braking surface, an inertia mass entrained by the other components, non-deformable and distinct brake actuating gear, independently mobile in relation to the other braking system and fixed and controlled by the inertia mass, positively to engage the braking surface. This damper is for retaining the parts of a nuclear power station so that can withstand earthquakes [fr

  11. Thermonuclear device

    International Nuclear Information System (INIS)

    Inoue, Toyokazu; Murata, Toru.

    1983-01-01

    Purpose: To shield superconducting coils for use in the generation of magnetic field against neutron irradiation thereby preventing tritium contamination. Constitution: The thermonuclear device comprises, in its inside, a vacuum container for containing plasmas, superconducting coils disposed to the outside of the vacuum container and neutron absorbers disposed between the super-conducting coils and the vacuum container. since neutrons issued from the plasma are absorbed by neutron absorbers and not irradiated to the superconducting coils, generation of tritium due to the reaction between 3 He in the liquid helium as the coolants for the super-conducting coils and the neutrons is prevented. (Aizawa, K.)

  12. Current status of rotational atherectomy.

    Science.gov (United States)

    Tomey, Matthew I; Kini, Annapoorna S; Sharma, Samin K

    2014-04-01

    Rotational atherectomy facilitates percutaneous coronary intervention for complex de novo lesions with severe calcification. A strategy of routine rotational atherectomy has not, however, conferred reduction in restenosis or major adverse cardiac events. As it is technically demanding, rotational atherectomy is also uncommon. At this 25-year anniversary since the introduction of rotational atherectomy, we sought to review the current state-of-the-art in rotational atherectomy technique, safety, and efficacy data in the modern era of drug-eluting stents, strategies to prevent and manage complications, including slow-flow/no-reflow and burr entrapment, and appropriate use in the context of the broader evolution in the management of stable ischemic heart disease. Fundamental elements of optimal technique include use of a single burr with burr-to-artery ratio of 0.5 to 0.6-rotational speed of 140,000 to 150,000 rpm, gradual burr advancement using a pecking motion, short ablation runs of 15 to 20 s, and avoidance of decelerations >5,000 rpm. Combined with meticulous technique, optimal antiplatelet therapy, vasodilators, flush solution, and provisional use of atropine, temporary pacing, vasopressors, and mechanical support may prevent slow-flow/no-reflow, which in contemporary series is reported in 0.0% to 2.6% of cases. On the basis of the results of recent large clinical trials, a subset of patients with complex coronary artery disease previously assigned to rotational atherectomy may be directed instead to medical therapy alone or bypass surgery. For patients with de novo severely calcified lesions for which rotational atherectomy remains appropriate, referral centers of excellence are required. Copyright © 2014 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  13. Rotating Shadowband Spectroradiometer (RSS) Handbook

    Energy Technology Data Exchange (ETDEWEB)

    Kiedron, P; Schlemmer, J; Klassen, M

    2005-01-01

    The rotating shawdowband spectroradiometer (RSS) implements the same automated shadowbanding technique used by the multifilter rotating shadowband radiometer (MFRSR), and so it too provides spectrally-resolved, direct-normal, diffuse-horizontal, and total-horizontal irradiances, and can be calibrated in situ via Langley regression. The irradiance spectra are measured simultaneously at all spectral elements (pixels) in 360-nm to 1050-nm range.

  14. Faraday rotation measurements at Ootacamund

    Science.gov (United States)

    Sethia, G.; Chandra, H.; Deshpande, M. R.; Rastogi, R. G.

    1978-01-01

    The results of Faraday rotation measurements made at Ootacamund during ATS-6 phase II are presented. For summer and equinoctial months, even though no clear noon bite-out is observed in the variation of Faraday a decrease is observed in the rate of increase of rotation around 0900-1000 hours LT. This is attributed to the 'fountain effect' which is responsible for the noontime bite-out in F2-region peak electron density.

  15. Power calculation of grading device in desintegrator

    Science.gov (United States)

    Bogdanov, V. S.; Semikopenko, I. A.; Vavilov, D. V.

    2018-03-01

    This article describes the analytical method of measuring the secondary power consumption, necessitated by the installation of a grading device in the peripheral part of the grinding chamber in the desintegrator. There is a calculation model for defining the power input of the disintegrator increased by the extra power demand, required to rotate the grading device and to grind the material in the area between the external row of hammers and the grading device. The work has determined the inertia moments of a cylindrical section of the grading device with armour plates. The processing capacity of the grading device is adjusted to the conveying capacity of the auger feeder. The grading device enables one to increase the concentration of particles in the peripheral part of the grinding chamber and the amount of interaction between particles and armour plates as well as the number of colliding particles. The perforated sections provide the output of the ground material with the proper size granules, which together with the effects of armour plates, improves the efficiency of grinding. The power demand to rotate the grading device does not exceed the admissible value.

  16. Rotating carbon nanotube membrane filter for water desalination

    Science.gov (United States)

    Tu, Qingsong; Yang, Qiang; Wang, Hualin; Li, Shaofan

    2016-01-01

    We have designed a porous nanofluidic desalination device, a rotating carbon nanotube membrane filter (RCNT-MF), for the reverse osmosis desalination that can turn salt water into fresh water. The concept as well as design strategy of RCNT-MF is modeled, and demonstrated by using molecular dynamics simulation. It has been shown that the RCNT-MF device may significantly improve desalination efficiency by combining the centrifugal force propelled reverse osmosis process and the porous CNT-based fine scale selective separation technology. PMID:27188982

  17. A paper-polymer centrifugal device for low-cost sample pre-concentration and colorimetric lateral flow assay enhancement

    CSIR Research Space (South Africa)

    Wiederoder, MS

    2016-10-01

    Full Text Available This study describes a novel hybrid paper-polymer centrifugal microfluidic device for pre-concentration of E.coli and lateral flow immunoassay enhancement for water quality verification. The device balances rotational centrifugal force...

  18. Relaxation processes in rotational motion

    International Nuclear Information System (INIS)

    Broglia, R.A.

    1986-01-01

    At few MeV above the yrast line the normally strong correlations among γ-ray energies in a rotational sequence become weaker. This observation can be interpreted as evidence for the damping of rotational motion in hot nuclei. It seems possible to relate the spreading width of the E2-rotational decay strength to the spread in frequency Δω 0 of rotational bands. The origin of these fluctuations is found in: (1) fluctuations in the occupation of special single-particle orbits which contribute a significant part of the total angular momentum; and (2) fluctuations in the moment of inertia induced by vibrations of the nuclear shape. Estimates of Δω 0 done making use of the hundred-odd known discrete rotational bands in the rare-earth region lead, for moderate spin and excitation energies (I ≅ 30 and U ≅ 3 to 4 MeV), to rotational spreading widths of the order of 60 to 160 keV in overall agreement with the data. 24 refs

  19. The rotation of galaxy clusters

    International Nuclear Information System (INIS)

    Tovmassian, H.M.

    2015-01-01

    The method for detection of the galaxy cluster rotation based on the study of distribution of member galaxies with velocities lower and higher of the cluster mean velocity over the cluster image is proposed. The search for rotation is made for flat clusters with a/b> 1.8 and BMI type clusters which are expected to be rotating. For comparison there were studied also round clusters and clusters of NBMI type, the second by brightness galaxy in which does not differ significantly from the cluster cD galaxy. Seventeen out of studied 65 clusters are found to be rotating. It was found that the detection rate is sufficiently high for flat clusters, over 60 per cent, and clusters of BMI type with dominant cD galaxy, ≈ 35 per cent. The obtained results show that clusters were formed from the huge primordial gas clouds and preserved the rotation of the primordial clouds, unless they did not have mergings with other clusters and groups of galaxies, in the result of which the rotation has been prevented

  20. HINT computation of LHD equilibrium with zero rotational transform surface

    International Nuclear Information System (INIS)

    Kanno, Ryutaro; Toi, Kazuo; Watanabe, Kiyomasa; Hayashi, Takaya; Miura, Hideaki; Nakajima, Noriyoshi; Okamoto Masao

    2004-01-01

    A Large Helical Device equilibrium having a zero rotational transform surface is studied by using the three dimensional MHD equilibrium code, HINT. We find existence of the equilibrium but with formation of the two or three n=0 islands composing a homoclinic-type structure near the center, where n is a toroidal mode number. The LHD equilibrium maintains the structure, when the equilibrium beta increases. (author)

  1. Scalable devices

    KAUST Repository

    Krüger, Jens J.

    2014-01-01

    In computer science in general and in particular the field of high performance computing and supercomputing the term scalable plays an important role. It indicates that a piece of hardware, a concept, an algorithm, or an entire system scales with the size of the problem, i.e., it can not only be used in a very specific setting but it\\'s applicable for a wide range of problems. From small scenarios to possibly very large settings. In this spirit, there exist a number of fixed areas of research on scalability. There are works on scalable algorithms, scalable architectures but what are scalable devices? In the context of this chapter, we are interested in a whole range of display devices, ranging from small scale hardware such as tablet computers, pads, smart-phones etc. up to large tiled display walls. What interests us mostly is not so much the hardware setup but mostly the visualization algorithms behind these display systems that scale from your average smart phone up to the largest gigapixel display walls.

  2. Thermonuclear device

    International Nuclear Information System (INIS)

    Honda, Takuro; Maki, Koichi.

    1997-01-01

    The present invention provides a thermonuclear device, in which integrity of a measuring device is kept, the reactor wall temperature and wear of armour materials are monitored accurately even under intense radiation rays, so that the flow rate of coolants and plasma power can be controlled by using the signals. Infrared rays generated from the surface of the armour materials disposed on a first wall are detected to measure the reactor wall temperature. Coolant flow rate and plasma power are controlled based on the obtained reactor wall temperature. In addition, infrared rays generated from the back of the armour materials are detected to obtain the surface temperature in order to avoid intense radiation rays from plasmas. The coolant flow rate and the plasma power are controlled based on the obtained temperature on the surface of the reactor thereby controlling the temperature of the first wall and the armour material to 300degC or lower in a case of the first wall made of stainless steel and 1000degC or lower in a case of the armour material made of graphite. (I.S.)

  3. Thermonuclear device

    International Nuclear Information System (INIS)

    Kuriyama, Masaaki; Yamamoto, Masahiro; Furuyama, Masayuki; Saito, Ryusei.

    1981-01-01

    Purpose: To enable the efficient and rapid cooling of a vacuum vessel by cooling with gas when the temperature of the vacuum vessel is higher than the boiling point of water and cooling with water when the temperature is lower than the boiling point of water. Constitution: A cooling pipe is provided through an insulating pipe on the outer periphery of a vacuum vessel. The cooling pipe communicates through a cooling gas valve and a coolant valve with a cooling gas supply device and a coolant supply device, and a heat exchanger is disposed at the pipe. When the vessel is higher than the boiling point of the coolant the coolant valve is closed and the cooling gas valve is opened and gas is supplied to cool the vessel. The gas is recoverd through a heat exchanger. On the other hand, when the temperature of vessel is lower than the boiling point of the coolant, the gas valve is closed, the coolant valve is opened, and the vessel is cooled with coolant. The vacuum vessel can be cooled for short time employing both the gas and the coolant together. (Yoshino, Y.)

  4. Thermonuclear device

    International Nuclear Information System (INIS)

    Takano, Hirohisa; Nakamoto, Kazunari; Hanai, Satoshi.

    1984-01-01

    Purpose: To provide coils of high mechanical strength for use at the center of a torus type thermonuclear device. Constitution: A plurality of copper plates having cooling holes and bolt holes and insulation paper sheets of the same shape are prepared. The copper plate is different from the insulation paper sheet only in that the position-phase angle of the opening portion is larger by 15 - 30 0 . The copper plates and the insulation paper sheets are alternately stacked by a required number of turns while displacing the angle, and then clamped by bolts to form a mechanically strong coil with no metallurgical joining. Further, since the insulation paper sheets are not present in the radial direction and only one insulation paper sheet is inserted for each turn in the direction of the coil height, the space occupied by the coil can be decreased. According to this invention, the magnetic flux density at the center of the device can be increased as compared with the conventional case to thereby apply a higher voltage on the side of plasmas. (Moriyama, K.)

  5. Irradiation device

    International Nuclear Information System (INIS)

    Suzuki, Toshimitsu.

    1989-01-01

    In an irradiation device for irradiating radiation rays such as electron beams to pharmaceuticals, etc., since the distribution of scanned electron rays was not monitored, the electron beam intensity could be determined only indirectly and irradiation reliability was not satisfactory. In view of the above, a plurality of monitor wires emitting secondary electrons are disposed in the scanning direction near a beam take-out window of a scanning duct, signals from the monitor wires are inputted into a display device such as a cathode ray tube, as well as signals from the monitor wires at the central portion are inputted into counting rate meters to measure the radiation dose as well. Since secondary electrons are emitted when electron beams pass through the monitor wires and the intensity thereof is in proportion with the intensity of incident electron beams, the distribution of the radiation dose can be monitored by measuring the intensity of the emitted secondary electrons. Further, uneven irradiation, etc. can also be monitored to make the radiation of irradiation rays reliable. (N.H.)

  6. The global percutaneous shuttling technique tip for arthroscopic rotator cuff repair

    Directory of Open Access Journals (Sweden)

    Bryan G. Vopat

    2014-05-01

    Full Text Available Most arthroscopic rotator cuff repairs utilize suture passing devices placed through arthro- scopic cannulas. These devices are limited by the size of the passing device where the suture is passed through the tendon. An alternative technique has been used in the senior author’s practice for the past ten years, where sutures are placed through the rotator cuff tendon using percutaneous passing devices. This technique, dubbed the global percutaneous shuttling technique of rotator cuff repair, affords the placement of sutures from nearly any angle and location in the shoulder, and has the potential advantage of larger suture bites through the tendon edge. These advantages may increase the area of tendon available to compress to the rotator cuff footprint and improve tendon healing and outcomes. The aim of this study is to describe the global percutaneous shuttling (GPS technique and report our results using this method. The GPS technique can be used for any full thickness rotator cuff tear and is particularly useful for massive cuff tears with poor tissue quality. We recently followed up 22 patients with an average follow up of 32 months to validate its usefulness. American Shoulder and Elbow Surgeons scores improved significantly from 37 preoperatively to 90 postoperatively (P<0.0001. This data supports the use of the GPS technique for arthroscopic rotator cuff repair. Further biomechanical studies are currently being performed to assess the improvements in tendon footprint area with this technique.

  7. Open source laboratory sample rotator mixer and shaker

    Directory of Open Access Journals (Sweden)

    Karankumar C. Dhankani

    2017-04-01

    Full Text Available An open-source 3-D printable laboratory sample rotator mixer is developed here in two variants that allow users to opt for the level of functionality, cost saving and associated complexity needed in their laboratories. First, a laboratory sample rotator is designed and demonstrated that can be used for tumbling as well as gentle mixing of samples in a variety of tube sizes by mixing them horizontally, vertically, or any position in between. Changing the mixing angle is fast and convenient and requires no tools. This device is battery powered and can be easily transported to operate in various locations in a lab including desktops, benches, clean hoods, chemical hoods, cold rooms, glove boxes, incubators or biological hoods. Second, an on-board Arduino-based microcontroller is incorporated that adds the functionality of a laboratory sample shaker. These devices can be customized both mechanically and functionally as the user can simply select the operation mode on the switch or alter the code to perform custom experiments. The open source laboratory sample rotator mixer can be built by non-specialists for under US$30 and adding shaking functionality can be done for under $20 more. Thus, these open source devices are technically superior to the proprietary commercial equipment available on the market while saving over 90% of the costs.

  8. Practical microwave electron devices

    CERN Document Server

    Meurant, Gerard

    2013-01-01

    Practical Microwave Electron Devices provides an understanding of microwave electron devices and their applications. All areas of microwave electron devices are covered. These include microwave solid-state devices, including popular microwave transistors and both passive and active diodes; quantum electron devices; thermionic devices (including relativistic thermionic devices); and ferrimagnetic electron devices. The design of each of these devices is discussed as well as their applications, including oscillation, amplification, switching, modulation, demodulation, and parametric interactions.

  9. Controllable rotational inversion in nanostructures with dual chirality.

    Science.gov (United States)

    Dai, Lu; Zhu, Ka-Di; Shen, Wenzhong; Huang, Xiaojiang; Zhang, Li; Goriely, Alain

    2018-04-05

    Chiral structures play an important role in natural sciences due to their great variety and potential applications. A perversion connecting two helices with opposite chirality creates a dual-chirality helical structure. In this paper, we develop a novel model to explore quantitatively the mechanical behavior of normal, binormal and transversely isotropic helical structures with dual chirality and apply these ideas to known nanostructures. It is found that both direction and amplitude of rotation can be finely controlled by designing the cross-sectional shape. A peculiar rotational inversion of overwinding followed by unwinding, observed in some gourd and cucumber tendril perversions, not only exists in transversely isotropic dual-chirality helical nanobelts, but also in the binormal/normal ones when the cross-sectional aspect ratio is close to 1. Beyond this rotational inversion region, the binormal and normal dual-chirality helical nanobelts exhibit a fixed directional rotation of unwinding and overwinding, respectively. Moreover, in the binormal case, the rotation of these helical nanobelts is nearly linear, which is promising as a possible design for linear-to-rotary motion converters. The present work suggests new designs for nanoscale devices.

  10. METHOD FOR DETERMINATION OF ROTATION CENTER IN VIBRATING OBJECT

    Directory of Open Access Journals (Sweden)

    I. P. Kauryha

    2016-01-01

    Full Text Available Linear piezoelectric gauges, eddy current transducers and other control and measuring devices have been widely applied for vibration diagnostics of objects in industry. Methods based on such gauges and used for measuring angular and linear vibrations do not provide the possibility to assess a rotation center or point angle of an object. Parasitic oscillations may occur during rotor rotation and in some cases the oscillations are caused by dis-balance. The known methods for measuring angular and linear vibrations make it possible to detect the phenomenon and they do not provide information for balancing of the given object. For this very reason the paper describes a method for obtaining instantaneous rotation center in the vibrating object. It allows to improve informational content of the measurements owing to obtaining additional data on position of object rotation center. The obtained data can be used for balancing of a control object. Essence of the given method is shown by an example of piezoelectric gauges of linear vibrations. Two three-axial gauges are fixed to the investigated object. Then gauge output signals are recalculated in angular vibrations of the object (for this purpose it is necessary to know a distance between gauges. Further projection positions of the object rotation center are determined on three orthogonal planes. Instantaneous rotation center is calculated according to the position of one of the gauges. The proposed method permits to obtain data on linear and angular vibrations and rotation center position of the vibrating object using one system of linear gauge. Possibilities of object diagnostics are expanded due to increase in number of determined parameters pertaining to object moving. The method also makes it possible to reduce material and time expenses for measurement of an angular vibration component. 

  11. Frequency Shift of a Rotating Mass-Imbalance Immersed in an Acoustic Fluid

    International Nuclear Information System (INIS)

    Stephen R. Novascone; David M. Weinberg; Michael J. Anderson

    2005-01-01

    In this paper, we describe a physical mechanism that relates a measurable behavior of a vibrating device to the physical properties of a surrounding acoustic medium. The vibrating device under consideration is a rotating imbalance immersed in an unbounded acoustic fluid. It is assumed that the rotating imbalance is driven by an electromagnetic motor excited by a given DC voltage. If nonlinearities are ignored, the steady state operational frequency of such a device is determined by a balance between the applied electromagnetic and opposing frictional torque on the rotating imbalance. If nonlinearities are retained, it is shown that under certain circumstances, the surrounding acoustic medium exerts an additional time-averaged opposing torque on the rotating imbalance that reduces the operational frequency of the device. Consequently, the operational frequency of the device becomes linked to the physical properties of the surrounding medium. Analytical calculations showed that the radiative resistance of an acoustic fluid caused the opposing torque. The shift in frequency is proportional to the radiative resistance and the square of the rotating eccentricity, but inversely proportional the total transducer mass and the damping effect of the DC motor

  12. Thermonuclear device

    International Nuclear Information System (INIS)

    Furuyama, Masayuki.

    1979-01-01

    Purpose: To provide the subject device wherein a conductive short-circuiting ring is installed in the vicinity of the bonded part of bellows and thick portion of vacuum vessel in the small circumferential direction of torus, thereby to reduce the electromagnetic force generated at the bellows. Constitution: A conductive short-circuiting ring is provided in the vicinity of the connected part of a thick portion and bellows portion. By this organization, a saddle type current generated at the thick portion by a vertical magnetic field flows through the short-circuiting ring because the resistance at a part where the short-circuiting ring is provided is reduced, and the current flowing through the bellows is remarkably reduced. For this reason, the electromagnetic force generated at the bellows is reduced thereby to prevent the bellows from being destroyed by the electromagnetic force. (Yoshihara, H.)

  13. Electrophoresis device

    Science.gov (United States)

    Rhodes, P. H.; Snyder, R. S. (Inventor)

    1982-01-01

    A device for separating cellular particles of a sample substance into fractionated streams of different cellular species includes a casing having a distribution chamber, a separation chamber, and a collection chamber. The electrode chambers are separated from the separation chamber interior by means of passages such that flow variations and membrane variations around the slotted portion of the electrode chamber do not enduce flow perturbations into the laminar buffer curtain flowing in the separation chamber. The cellular particles of the sample are separated under the influence of the electrical field and the separation chamber into streams of different cellular species. The streams of separated cells enter a partition array in the collection chamber where they are fractionated and collected.

  14. Thermonuclear device

    International Nuclear Information System (INIS)

    Suzuki, Shohei

    1988-01-01

    Purpose: To obtain high voltage withstanding current introduction terminals not suffering from the effects of the reduction in the creeping voltage withstanding property by the application of magnetic fields. Constitution: This invention concerns a current introduction terminal for supplying electric current to coils for use in a thermonuclear device, etc. The conductor of the current introduction terminal on the side of vacuum is completely covered with solid insulator. This can eliminate the portion of securing the creeping withstanding voltage. The voltage withstanding characteristics of the solid insulator covering the portion of the conductor on the side of vacuum has a constant value irrespective of the atmosphere or the absence or presence of magnetic fields. Accordingly, the voltage withstanding characteristics of the current introduction terminal on the side of vacuum are determined by the property of the solid insulator, which is not reduced by the application of magnetic fields. (Ikeda, J.)

  15. Design for coordinated measurements of Faraday rotation and line-of-sight electron density using heterodyne techniques

    International Nuclear Information System (INIS)

    Jacobson, A.R.

    1977-07-01

    This report proposes a device which can overcome certain of the compromises of conventional Faraday rotation methods and at the same time measure the optical phase as well as the polarization. This would be useful for unfolding the Faraday rotation signal using the line-of-sight density along exactly the same path. Preliminary design parameters using a CO 2 laser are presented

  16. Safety device of thermonuclear device

    International Nuclear Information System (INIS)

    Aoki, Isao; Ueda, Shuzo; Seki, Yasushi; Sakurai, Akiko; Kasahara, Fumio; Obara, Atsushi; Yamauchi, Michinori.

    1997-01-01

    The present invention provides a safety device against an event of intrusion of coolants in a vacuum vessel. Namely, a coolant supply system comprises cooling tubes for supplying coolants to main reactor structure components including a vacuum vessel. A detection means detects leakage of coolants in the vacuum vessel. A coolant supply control means controls the supply of coolants to the main reactor structural components based on the leakage detection signals of the detection means. A stagnated material discharging means discharges stagnated materials in the main reactor structural components caused by the leakage of coolants. The leakage of coolants (for example, water) in the vacuum vessel can thus be detected by the water detection device in the vacuum vessel. A control value of a coolant supply means is closed by the leakage detection signals. The supply of coolants to the main reactor structural components is restricted to suppress the leakage. The stagnated materials are discharged to a tank by way of a water draining valve. (I.S.)

  17. RFQ device for accelerating particles

    Science.gov (United States)

    Shepard, K.W.; Delayen, J.R.

    1995-06-06

    A superconducting radio frequency quadrupole (RFQ) device includes four spaced elongated, linear, tubular rods disposed parallel to a charged particle beam axis, with each rod supported by two spaced tubular posts oriented radially with respect to the beam axis. The rod and post geometry of the device has four-fold rotation symmetry, lowers the frequency of the quadrupole mode below that of the dipole mode, and provides large dipole-quadrupole mode isolation to accommodate a range of mechanical tolerances. The simplicity of the geometry of the structure, which can be formed by joining eight simple T-sections, provides a high degree of mechanical stability, is insensitive to mechanical displacement, and is particularly adapted for fabrication with superconducting materials such as niobium. 5 figs.

  18. Waste classifying and separation device

    International Nuclear Information System (INIS)

    Kakiuchi, Hiroki.

    1997-01-01

    A flexible plastic bags containing solid wastes of indefinite shape is broken and the wastes are classified. The bag cutting-portion of the device has an ultrasonic-type or a heater-type cutting means, and the cutting means moves in parallel with the transferring direction of the plastic bags. A classification portion separates and discriminates the plastic bag from the contents and conducts classification while rotating a classification table. Accordingly, the plastic bag containing solids of indefinite shape can be broken and classification can be conducted efficiently and reliably. The device of the present invention has a simple structure which requires small installation space and enables easy maintenance. (T.M.)

  19. RFQ device for accelerating particles

    Energy Technology Data Exchange (ETDEWEB)

    Shepard, Kenneth W. (Park Ridge, IL); Delayen, Jean R. (Naperville, IL)

    1995-01-01

    A superconducting radio frequency quadrupole (RFQ) device includes four spaced elongated, linear, tubular rods disposed parallel to a charged particle beam axis, with each rod supported by two spaced tubular posts oriented radially with respect to the beam axis. The rod and post geometry of the device has four-fold rotation symmetry, lowers the frequency of the quadrupole mode below that of the dipole mode, and provides large dipole-quadrupole mode isolation to accommodate a range of mechanical tolerances. The simplicity of the geometry of the structure, which can be formed by joining eight simple T-sections, provides a high degree of mechanical stability, is insensitive to mechanical displacement, and is particularly adapted for fabrication with superconducting materials such as niobium.

  20. Dynamic Models of Instruments Using Rotating Unbalanced Masses

    Science.gov (United States)

    Hung, John Y.; Gallaspy, Jason M.; Bishop, Carlee A.

    1998-01-01

    The motion of telescopes, satellites, and other flight bodies have been controlled by various means in the past. For example, gimbal mounted devices can use electric motors to produce pointing and scanning motions. Reaction wheels, control moment gyros, and propellant-charged reaction jets are other technologies that have also been used. Each of these methods has its advantages, but all actuator systems used in a flight environment face the challenges of minimizing weight, reducing energy consumption, and maximizing reliability. Recently, Polites invented and patented the Rotating Unbalanced Mass (RUM) device as a means for generation scanning motion on flight experiments. RUM devices together with traditional servomechanisms have been successfully used to generate various scanning motions: linear, raster, and circular. The basic principle can be described: A RUM rotating at constant angular velocity exerts a cyclic centrifugal force on the instrument or main body, thus producing a periodic scanning motion. A system of RUM devices exerts no reaction forces on the main body, requires very little energy to rotate the RUMS, and is simple to construct. These are significant advantages over electric motors, reaction wheels, and control moment gyroscopes. Although the RUM device very easily produces scanning motion, an auxiliary control system has been required to maintain the proper orientation, or pointing of the main body. It has been suggested that RUM devices can be used to control pointing dynamics, as well as generate the desired periodic scanning motion. The idea is that the RUM velocity will not be kept constant, but will vary over the period of one RUM rotation. The thought is that the changing angular velocity produces a centrifugal force having time-varying magnitude and direction. The scope of this ongoing research project is to study the pointing control concept, and recommend a direction of study for advanced pointing control using only RUM devices. This

  1. Refueling system with small diameter rotatable plugs

    International Nuclear Information System (INIS)

    Ritz, W.C.

    1987-01-01

    This patent describes a liquid-metal fastbreeder nuclear reactor comprising a reactor pressure vessel and closure head therefor, a reactor core barrel disposed within the reactor vessel and enclosing a reactor core having therein a large number of closely spaced fuel assemblies, and the reactor core barrel and the reactor core having an approximately concentric circular cross-sectional configuration with a geometric center in predetermined location within the reactor vessel. The improved refueling system described here comprises: a large controllably rotatable plug means comprising the substantial portion of the closure head, a reactor upper internals structure mounted from the large rotatable plug means. The large rotatable plug means has an approximately circular configuration which approximates the cross-sectional configuration of the reactor core barrel with a center of rotation positioned a first predetermined distance from the geometric center of the reactor core barrel so that the large rotatable plug means rotates eccentrically with respect to the reactor core barrel; a small controllably rotatable plug means affixed to the large rotatable plug means and rotatable with respect thereto. The small rotatable plug means has a center of rotation which is offset a second predetermined distance from the rotational center of the large rotatable plug means so that the small rotatable plug means rotates eccentrically with respect to the large rotatable plug means

  2. A new concept for the control of a slow-extracted beam in a line with rotational optics: Part II

    CERN Document Server

    Benedikt, Michael; Pullia, M

    1999-01-01

    The current trend in hadrontherapy is towards high-precision, conformal scanning of tumours with a 'pencil' beam of light ions or protons, delivered by a synchrotron using slow extraction. The particular shape of the slow-extracted beam segment in phase space and the need to vary the beam size in a lattice with rotating optical elements create a special problem for the design of the extraction transfer line and gantry. The design concept presented in this report is based on telescope modules with integer-pi phase advances in both transverse planes. The beam size in the plane of the extraction is controlled by altering the phase advance and hence the rotation of the extracted beam segment in phase space. The vertical beam size is controlled by stepping the vertical betatron amplitude function over a range of values and passing the changed beam size from 'hand-to-hand' through the telescope modules to the various treatment rooms. In the example given, a combined phase shifter and 'stepper', at a point close to ...

  3. A new concept for the control of a slow-extracted beam in a line with rotational optics, 2

    CERN Document Server

    Benedikt, Michael; Pullia, M

    1999-01-01

    For pt.I see ibid., vol.430, p.512-22, 1999. The current trend in hadron therapy is towards high-precision, conformal scanning of tumours with a `pencil' beam of light ions or protons, delivered by a synchrotron using slow extraction. The particular shape of the slow- extracted beam segment in phase space and the need to vary the beam size in a lattice with rotating optical elements create a special problem for the design of the extraction transfer line and gantry. The design concept presented in this report is based on telescope modules with integer- pi phase advances in both transverse planes. The beam size in the plane of the extraction is controlled by altering the phase advance and hence the rotation of the extracted beam segment in phase space. The vertical beam size is controlled by stepping the vertical betatron amplitude function over a range of values and passing the changed beam size from `hand-to-hand' through the telescope modules to the various treatment rooms. In the example given, a combined p...

  4. A New Concept for the Control of a Slow-Extracted Beam in a Line with Rotational Optics, 2

    CERN Document Server

    Benedikt, Michael; Pullia, M

    1999-01-01

    The current trend in hadrontherapy is towards high-precision, conformal scanning of tumours with a 'pencil' beam of light ions, or protons, delivered by a synchrotron using slow-extraction. The particular shape of the slow-extracted beam segment in phase space and the need to vary the beam size in a lattice with rotating optical elements create a special problem for the design of the extraction transfer line and gantry. The design concept presented in this report is based on telescope modules with integer-p phase advances in both transverse planes. The beam size in the plane of the extraction is controlled by altering the phase advance and hence the rotation of the extracted beam segment in phase space. The vertical beam size is controlled by stepping the vertical betatron amplitude function over a range of values and passing the changed beam size from 'hand-to-hand' through the telescope modules to the various treatment rooms. In the example given, a combined phase-shifter and 'stepper', at a point close to ...

  5. Tokamak rotation and charge exchange

    International Nuclear Information System (INIS)

    Hazeltine, R.D.; Rowan, W.L.; Solano, E.R.; Valanju, P.M.

    1991-01-01

    In the absence of momentum input, tokamak toroidal rotation rates are typically small - no larger in particular than poloidal rotation - even when the radial electric field is strong, as near the plasma edge. This circumstance, contradicting conventional neoclassical theory, is commonly attributed to the rotation damping effect of charge exchange, although a detailed comparison between charge-exchange damping theory and experiment is apparently unavailable. Such a comparison is attempted here in the context of recent TEXT experiments, which compare rotation rates, both poloidal and toroidal, in helium and hydrogen discharges. The helium discharges provide useful data because they are nearly free of ion-neutral charge exchange; they have been found to rotate toroidally in reasonable agreement with neoclassical predictions. The hydrogen experiments show much smaller toroidal motion as usual. The theoretical calculation uses the full charge-exchange operator and assumes plateau collisionality, roughly consistent with the experimental conditions. The authors calculate the ion flow as a function of v cx /v c , where v cx is the charge exchange rate and v c the Coulomb collision frequency. The results are in reasonable accord with the observations. 1 ref

  6. Development of Rotational Accelerometers Final Report CRADA No. TSB-2008-99

    Energy Technology Data Exchange (ETDEWEB)

    Hunter, S. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States); Crosson, R. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2017-10-16

    One of the difficulties in fabricating an inexpensive angular rate or rotation sensor is producing a device that is insensitive to acceleration, including the constant acceleration of gravity. The majority of rate sensors are either tuning fork type devices sensing a relatively weak force (i.e., Coriolis effect) and thus not very sensitive, or gyroscopes (either rotating or fiber optic based) that are large, consume lots of power and are expensive. This project was a collaborative effort between LLNL and The Fredericks Company to develop a rotational sensor as a standardized, commercial product. The Fredericks Company possessed expertise and capabilities in the technical aspects of manufacturing this type of sensor, and they were interested in collaborating with LLNL to manufacture the rotational rate sensors as a commercial product.

  7. The structure of rotational discontinuities

    International Nuclear Information System (INIS)

    Neugebauer, M.

    1989-01-01

    This study examines the structures of a set of rotational discontinuities detected in the solar wind by the ISEE-3 spacecraft. It is found that the complexity of the structure increases as the angle θ between the propagation vector k and the magnetic field decreases. For rotational discontinuities that propagate at a large angle to the field with an ion (left-hand) sense of rotation, the magnetic hodograms tend to be flattened, in agreement with prior numerical simulations. When θ is large, angular overshoots are often observed at one or both ends of the discontinuity. When the propagation is nearly parallel to the field (i.e., when θ is small), many different types of structure are seen, ranging from straight lines, the S-shaped curves, to complex, disorganized shapes

  8. Motions on a rotating planet

    Science.gov (United States)

    Schröer, H.

    In chapter 1 we want to describe the motion of a falling body on a rotating planet. The planet rotates with an arbitrary changable angular velocity and has a translational acceleration. We obtain 3 differential equations. For the general gravitational field an exact solution is possible, when the differential equation system is explicit solvable. Then we consider the case, if the angular velocity and the translational acceleration is constant. With a special transformation we get 3 partial differential equations of first order. Instead of a planet sphere we can choose a general body of rotation. Even general bodies are possible. Chapter 2 contains the motion in a local coordinate system on planet's surface. We have an inhomogeneous linear differential equation of first order. If the angular velocity is constant, we get a system with constant coefficients. There is an english and a german edition.

  9. Thermonuclear device

    International Nuclear Information System (INIS)

    Suzuki, Shohei.

    1980-01-01

    Purpose: To improve the plasma confining efficiency in a thermonuclear device having magnet coils using super-conducting wires by decreasing the uneven magnetic field resulted from current supply terminals and wirings. Constitution: Current introduction terminals of magnet coils using superconducting wires are short circuitted with a superconducting short circuit wire. Upon supplying current to the coils, the resistance of the coils is rendered superconductive and the resistance of the short circuit wire is rendered normally conductive heated by a heater and the switch is closed. In this case, most parts of the current are flown through the resistance of the coils and the switch is opened when the current arrives at a predetermined value to render the resistance of the short circuit wire superconductive. Then, the current transfers from the thyristor power source to the resistance of the short circuit wire, whereby the resistance of the coils and that of the short circuit wire from a permanent current loop. In this conditions, since current flows through the short circuit wire and the coils and not to the current introduction terminals, no uniform magnetic field is generated. (Kawakami, Y.)

  10. Rotated balance in humans due to repetitive rotational movement

    Science.gov (United States)

    Zakynthinaki, M. S.; Madera Milla, J.; López Diaz De Durana, A.; Cordente Martínez, C. A.; Rodríguez Romo, G.; Sillero Quintana, M.; Sampedro Molinuevo, J.

    2010-03-01

    We show how asymmetries in the movement patterns during the process of regaining balance after perturbation from quiet stance can be modeled by a set of coupled vector fields for the derivative with respect to time of the angles between the resultant ground reaction forces and the vertical in the anteroposterior and mediolateral directions. In our model, which is an adaption of the model of Stirling and Zakynthinaki (2004), the critical curve, defining the set of maximum angles one can lean to and still correct to regain balance, can be rotated and skewed so as to model the effects of a repetitive training of a rotational movement pattern. For the purposes of our study a rotation and a skew matrix is applied to the critical curve of the model. We present here a linear stability analysis of the modified model, as well as a fit of the model to experimental data of two characteristic "asymmetric" elite athletes and to a "symmetric" elite athlete for comparison. The new adapted model has many uses not just in sport but also in rehabilitation, as many work place injuries are caused by excessive repetition of unaligned and rotational movement patterns.

  11. Apparatus and method for generating a magnetic field by rotation of a charge holding object

    Science.gov (United States)

    Gerald, II, Rex E.; Vukovic, Lela [Westchester, IL; Rathke, Jerome W [Homer Glenn, IL

    2009-10-13

    A device and a method for the production of a magnetic field using a Charge Holding Object that is mechanically rotated. In a preferred embodiment, a Charge Holding Object surrounding a sample rotates and subjects the sample to one or more magnetic fields. The one or more magnetic fields are used by NMR Electronics connected to an NMR Conductor positioned within the Charge Holding Object to perform NMR analysis of the sample.

  12. The rotate-plus-shift C-arm trajectory. Part I. Complete data with less than 180° rotation

    International Nuclear Information System (INIS)

    Ritschl, Ludwig; Fleischmann, Christof; Kuntz, Jan; Kachelrieß, Marc

    2016-01-01

    Purpose: In the last decade, C-arm-based cone-beam CT became a widely used modality for intraoperative imaging. Typically a C-arm CT scan is performed using a circular or elliptical trajectory around a region of interest. Therefore, an angular range of at least 180° plus fan angle must be covered to ensure a completely sampled data set. However, mobile C-arms designed with a focus on classical 2D applications like fluoroscopy may be limited to a mechanical rotation range of less than 180° to improve handling and usability. The method proposed in this paper allows for the acquisition of a fully sampled data set with a system limited to a mechanical rotation range of at least 180° minus fan angle using a new trajectory design. This enables CT like 3D imaging with a wide range of C-arm devices which are mainly designed for 2D imaging. Methods: The proposed trajectory extends the mechanical rotation range of the C-arm system with two additional linear shifts. Due to the divergent character of the fan-beam geometry, these two shifts lead to an additional angular range of half of the fan angle. Combining one shift at the beginning of the scan followed by a rotation and a second shift, the resulting rotate-plus-shift trajectory enables the acquisition of a completely sampled data set using only 180° minus fan angle of rotation. The shifts can be performed using, e.g., the two orthogonal positioning axes of a fully motorized C-arm system. The trajectory was evaluated in phantom and cadaver examinations using two prototype C-arm systems. Results: The proposed trajectory leads to reconstructions without limited angle artifacts. Compared to the limited angle reconstructions of 180° minus fan angle, image quality increased dramatically. Details in the rotate-plus-shift reconstructions were clearly depicted, whereas they are dominated by artifacts in the limited angle scan. Conclusions: The method proposed here employs 3D imaging using C-arms with less than 180° rotation

  13. The rotate-plus-shift C-arm trajectory. Part I. Complete data with less than 180° rotation

    Energy Technology Data Exchange (ETDEWEB)

    Ritschl, Ludwig; Fleischmann, Christof [Ziehm Imaging GmbH, Donaustraße 31, Nürnberg 90451 (Germany); Kuntz, Jan, E-mail: j.kuntz@dkfz.de; Kachelrieß, Marc [Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120 (Germany)

    2016-05-15

    Purpose: In the last decade, C-arm-based cone-beam CT became a widely used modality for intraoperative imaging. Typically a C-arm CT scan is performed using a circular or elliptical trajectory around a region of interest. Therefore, an angular range of at least 180° plus fan angle must be covered to ensure a completely sampled data set. However, mobile C-arms designed with a focus on classical 2D applications like fluoroscopy may be limited to a mechanical rotation range of less than 180° to improve handling and usability. The method proposed in this paper allows for the acquisition of a fully sampled data set with a system limited to a mechanical rotation range of at least 180° minus fan angle using a new trajectory design. This enables CT like 3D imaging with a wide range of C-arm devices which are mainly designed for 2D imaging. Methods: The proposed trajectory extends the mechanical rotation range of the C-arm system with two additional linear shifts. Due to the divergent character of the fan-beam geometry, these two shifts lead to an additional angular range of half of the fan angle. Combining one shift at the beginning of the scan followed by a rotation and a second shift, the resulting rotate-plus-shift trajectory enables the acquisition of a completely sampled data set using only 180° minus fan angle of rotation. The shifts can be performed using, e.g., the two orthogonal positioning axes of a fully motorized C-arm system. The trajectory was evaluated in phantom and cadaver examinations using two prototype C-arm systems. Results: The proposed trajectory leads to reconstructions without limited angle artifacts. Compared to the limited angle reconstructions of 180° minus fan angle, image quality increased dramatically. Details in the rotate-plus-shift reconstructions were clearly depicted, whereas they are dominated by artifacts in the limited angle scan. Conclusions: The method proposed here employs 3D imaging using C-arms with less than 180° rotation

  14. Protection device for a thermonuclear device

    International Nuclear Information System (INIS)

    Kawashima, Shuichi.

    1986-01-01

    Purpose: To exactly detect the void coefficients of coolants even under high magnetic fields thereby detect the overheat of a thermonuclear device at an early stage. Constitution: The protecting device of this invention comprises a laser beam generation device, a laser beam detection device and an accident detection device. The laser generation device always generates laser beams, which are permeated through coolants and detected by the laser beam detection device, the optical amount of which is transmitted to the accident detection device. The accident detection device judges the excess or insufficiency of the detected optical amount with respect to the optical amount of the laser beams under the stationary state as a reference and issues an accident signal. Since only the optical cables that do not undergo the effect of the magnetic fields are exposed to high magnetic fields in the protection device of this invention, a high reliability can be maintained. (Kamimura, M.)

  15. Mach's principle and rotating universes

    International Nuclear Information System (INIS)

    King, D.H.

    1990-01-01

    It is shown that the Bianchi 9 model universe satisfies the Mach principle. These closed rotating universes were previously thought to be counter-examples to the principle. The Mach principle is satisfied because the angular momentum of the rotating matter is compensated by the effective angular momentum of gravitational waves. A new formulation of the Mach principle is given that is based on the field theory interpretation of general relativity. Every closed universe with 3-sphere topology is shown to satisfy this formulation of the Mach principle. It is shown that the total angular momentum of the matter and gravitational waves in a closed 3-sphere topology universe is zero

  16. NH3 quantum rotators in Hofmann clathrates: intensity and width of rotational transition lines

    International Nuclear Information System (INIS)

    Vorderwisch, Peter; Sobolev, Oleg; Desmedt, Arnaud

    2004-01-01

    Inelastic structure factors for rotational transitions of uniaxial NH 3 quantum rotators, measured in a Hofmann clathrate with biphenyl as guest molecule, agree with those calculated for free rotators. A finite intrinsic line width, found for rotational transitions involving the rotational level j=3 at low temperature, supports a recently suggested model based on resonant rotor-rotor coupling

  17. Inertia thermonuclear device

    International Nuclear Information System (INIS)

    Madarame, Haruki; Nakamura, Norio; Oomura, Hiroshi.

    1983-01-01

    Purpose: To enable effective recovery of the thermonuclear reaction energy and effective protection of a cylinder metal against thermal destruction by forming a uniform and stable liquid metal wall to the inside of a cylindrical member. Constitution: Cylindrical body having a lateral axis is rotatably supported so that a liquid metal wall for use in the wet wall type thermonuclear device is formed centrifugally. A liquid metal injection port for injecting the liquid metal to the cylindrical member is disposed to the lateral axis and a liquid metal exit for flowing out the injected liquid metal is disposed to the body of the cylindrical member, so as to form a moving liquid metal layer flowing from the injection port through the inner circumferential surface of the cylindrical member to the liquid metal exit port. Then, the liquid metal is centrifugally forced to the inner surface of the cylindrical body to form a uniform and stable liquid metal wall at the inner surface of the cylindrical body, whereby the reaction energy can effectively be recovered and the cylinder metal can effectively be protected against thermal destruction. (Yoshihara, H.)

  18. Study on the Contra-Rotating Propeller system design and full-scale performance prediction method

    Directory of Open Access Journals (Sweden)

    Keh-Sik Min

    2009-09-01

    Full Text Available A ship's screw-propeller produces thrust by rotation and, at the same time, generates rotational flow behind the propeller. This rotational flow has no contribution to the generation of thrust, but instead produces energy loss. By recovering part of the lost energy in the rotational flow, therefore, it is possible to improve the propulsion efficiency. The contra-rotating propeller (CRP system is the representing example of such devices. Unfortunately, however, neither a design method nor a full-scale performance prediction procedure for the CRP system has been well established yet. The authors have long performed studies on the CRP system, and some of the results from the authors’ studies shall be presented and discussed.

  19. Methods of stabilizing a permanent maglev rotator in heart pumps and other rotary machines

    Directory of Open Access Journals (Sweden)

    kun-xi qian

    2014-10-01

    Full Text Available Permanent maglev rotator in a rotary machine could be stabilized according to the author’s experiences, by use of a non-PM (permanent magnetic force acting together with the PM force, and a non-PM bearing functioning together with the PM bearing, or a so-called gyro-effect which can stabilize all rotators including permanent maglev rotator. This paper presents both axially and radially driven permanent maglev centrifugal heart pumps, as well as a permanent maglev turbine machine and an industrially used permanent maglev centrifugal pump. In all this devices permanent maglev rotators achieve stable equilibrium by different approaches described in detail. Finally, the principle exhibition of gyro-effect and the route chart to stabilization of permanent maglev rotator are presented.

  20. Controllable nanoscale rotating actuator system based on carbon nanotube and graphene

    International Nuclear Information System (INIS)

    Huang, Jianzhang; Han, Qiang

    2016-01-01

    A controllable nanoscale rotating actuator system consisting of a double carbon nanotube and graphene driven by a temperature gradient is proposed, and its rotating dynamics performance and driving mechanism are investigated through molecular dynamics simulations. The outer tube exhibits stable pure rotation with certain orientation under temperature gradient and the steady rotational speed rises as the temperature gradient increases. It reveals that the driving torque is caused by the difference of atomic van der Waals potentials due to the temperature gradient and geometrical features of carbon nanotube. A theoretical model for driving torque is established based on lattice dynamics theory and its predicted results agree well with molecular dynamics simulations. Further discussion is taken according to the theoretical model. The work in this study would be a guide for design and application of controllable nanoscale rotating devices based on carbon nanotubes and graphene. (paper)

  1. Nerve–muscle activation by rotating permanent magnet configurations

    Science.gov (United States)

    Nicholson, Graham M.

    2016-01-01

    Key points The standard method of magnetic nerve activation using pulses of high current in coils has drawbacks of high cost, high electrical power (of order 1 kW), and limited repetition rate without liquid cooling.Here we report a new technique for nerve activation using high speed rotation of permanent magnet configurations, generating a sustained sinusoidal electric field using very low power (of order 10 W).A high ratio of the electric field gradient divided by frequency is shown to be the key indicator for nerve activation at high frequencies.Activation of the cane toad sciatic nerve and attached gastrocnemius muscle was observed at frequencies as low as 180 Hz for activation of the muscle directly and 230 Hz for curved nerves, but probably not in straight sections of nerve.These results, employing the first prototype device, suggest the opportunity for a new class of small low‐cost magnetic nerve and/or muscle stimulators. Abstract Conventional pulsed current systems for magnetic neurostimulation are large and expensive and have limited repetition rate because of overheating. Here we report a new technique for nerve activation, namely high‐speed rotation of a configuration of permanent magnets. Analytical solutions of the cable equation are derived for the oscillating electric field generated, which has amplitude proportional to the rotation speed. The prototype device built comprised a configuration of two cylindrical magnets with antiparallel magnetisations, made to rotate by interaction between the magnets’ own magnetic field and three‐phase currents in coils mounted on one side of the device. The electric field in a rectangular bath placed on top of the device was both numerically evaluated and measured. The ratio of the electric field gradient on frequency was approximately 1 V m−2 Hz−1 near the device. An exploratory series of physiological tests was conducted on the sciatic nerve and attached gastrocnemius muscle of the cane toad

  2. Nerve-muscle activation by rotating permanent magnet configurations.

    Science.gov (United States)

    Watterson, Peter A; Nicholson, Graham M

    2016-04-01

    The standard method of magnetic nerve activation using pulses of high current in coils has drawbacks of high cost, high electrical power (of order 1 kW), and limited repetition rate without liquid cooling. Here we report a new technique for nerve activation using high speed rotation of permanent magnet configurations, generating a sustained sinusoidal electric field using very low power (of order 10 W). A high ratio of the electric field gradient divided by frequency is shown to be the key indicator for nerve activation at high frequencies. Activation of the cane toad sciatic nerve and attached gastrocnemius muscle was observed at frequencies as low as 180 Hz for activation of the muscle directly and 230 Hz for curved nerves, but probably not in straight sections of nerve. These results, employing the first prototype device, suggest the opportunity for a new class of small low-cost magnetic nerve and/or muscle stimulators. Conventional pulsed current systems for magnetic neurostimulation are large and expensive and have limited repetition rate because of overheating. Here we report a new technique for nerve activation, namely high-speed rotation of a configuration of permanent magnets. Analytical solutions of the cable equation are derived for the oscillating electric field generated, which has amplitude proportional to the rotation speed. The prototype device built comprised a configuration of two cylindrical magnets with antiparallel magnetisations, made to rotate by interaction between the magnets' own magnetic field and three-phase currents in coils mounted on one side of the device. The electric field in a rectangular bath placed on top of the device was both numerically evaluated and measured. The ratio of the electric field gradient on frequency was approximately 1 V m(-2) Hz(-1) near the device. An exploratory series of physiological tests was conducted on the sciatic nerve and attached gastrocnemius muscle of the cane toad (Bufo marinus). Activation was

  3. Three-dimensional rotational plasma flows near solid surfaces in an axial magnetic field

    Energy Technology Data Exchange (ETDEWEB)

    Gorshunov, N. M., E-mail: gorshunov-nm@nrcki.ru; Potanin, E. P., E-mail: potanin45@yandex.ru [National Research Center Kurchatov Institute (Russian Federation)

    2016-11-15

    A rotational flow of a conducting viscous medium near an extended dielectric disk in a uniform axial magnetic field is analyzed in the magnetohydrodynamic (MHD) approach. An analytical solution to the system of nonlinear differential MHD equations of motion in the boundary layer for the general case of different rotation velocities of the disk and medium is obtained using a modified Slezkin–Targ method. A particular case of a medium rotating near a stationary disk imitating the end surface of a laboratory device is considered. The characteristics of a hydrodynamic flow near the disk surface are calculated within the model of a finite-thickness boundary layer. The influence of the magnetic field on the intensity of the secondary flow is studied. Calculations are performed for a weakly ionized dense plasma flow without allowance for the Hall effect and plasma compressibility. An MHD flow in a rotating cylinder bounded from above by a retarding cap is considered. The results obtained can be used to estimate the influence of the end surfaces on the main azimuthal flow, as well as the intensities of circulating flows in various devices with rotating plasmas, in particular, in plasma centrifuges and laboratory devices designed to study instabilities of rotating plasmas.

  4. Spectroscopy of Molecules in Extreme Rotational States Using AN Optical Centrifuge

    Science.gov (United States)

    Mullin, Amy S.; Toro, Carlos; Echibiri, Geraldine; Liu, Qingnan

    2012-06-01

    Our lab has developed a high-power optical centrifuge that is capable of trapping and spinning large number densities of molecules into extreme rotational states. By coupling this device with high resolution transient IR absorption spectroscopy, we measure the time-resolved behavior and energy profiles of individual ro-vibrational states of molecules in very high rotational states. Recent results will be discussed on the spectroscopy of new rotational states, collisional dynamics in the optical centrifuge, spatially-dependent energy profiles and possibilities for new chemistry induced by centrifugal forces.

  5. Numerical analysis on the action of centrifuge force in magnetic fluid rotating shaft seals

    Science.gov (United States)

    Zou, Jibin; Li, Xuehui; Lu, Yongping; Hu, Jianhui

    2002-11-01

    The magnetic fluid seal is suitable for high-speed rotating shaft seal applications. Centrifuge force will have evident influence on magnetic fluid rotating shaft seals. The seal capacity of the rotating shaft seal can be improved or increased by some measures. Through hydrodynamic analysis the moving status of the magnetic fluid is worked out. By numerical method, the magnetic field and the isobars in the magnetic fluid of a seal device are computed. Then the influence of the centrifuge force on the magnetic fluid seal is calculated quantitatively.

  6. Numerical analysis on the action of centrifuge force in magnetic fluid rotating shaft seals

    International Nuclear Information System (INIS)

    Zou Jibin; Li Xuehui; Lu Yongping; Hu Jianhui

    2002-01-01

    The magnetic fluid seal is suitable for high-speed rotating shaft seal applications. Centrifuge force will have evident influence on magnetic fluid rotating shaft seals. The seal capacity of the rotating shaft seal can be improved or increased by some measures. Through hydrodynamic analysis the moving status of the magnetic fluid is worked out. By numerical method, the magnetic field and the isobars in the magnetic fluid of a seal device are computed. Then the influence of the centrifuge force on the magnetic fluid seal is calculated quantitatively

  7. Surface micromachined counter-meshing gears discrimination device

    Science.gov (United States)

    Polosky, Marc A.; Garcia, Ernest J.; Allen, James J.

    2000-12-12

    A surface micromachined Counter-Meshing Gears (CMG) discrimination device which functions as a mechanically coded lock. Each of two CMG has a first portion of its perimeter devoted to continuous driving teeth that mesh with respective pinion gears. Each EMG also has a second portion of its perimeter devoted to regularly spaced discrimination gear teeth that extend outwardly on at least one of three levels of the CMG. The discrimination gear teeth are designed so as to pass each other without interference only if the correct sequence of partial rotations of the CMG occurs in response to a coded series of rotations from the pinion gears. A 24 bit code is normally input to unlock the device. Once unlocked, the device provides a path for an energy or information signal to pass through the device. The device is designed to immediately lock up if any portion of the 24 bit code is incorrect.

  8. Rotation of White Dwarf Stars

    OpenAIRE

    Kawaler, Steven D.

    2014-01-01

    I discuss and consider the status of observational determinations of the rotation velocities of white dwarf stars via asteroseismology and spectroscopy. While these observations have important implications on our understanding of the angular momentum evolution of stars in their late stages of evolution, more direct methods are sorely needed to disentangle ambiguities.

  9. Rotational damping motion in nuclei

    International Nuclear Information System (INIS)

    Egido, J.L.; Faessler, A.

    1991-01-01

    The recently proposed model to explain the mechanism of the rotational motion damping in nuclei is exactly solved. When compared with the earlier approximative solution, we find significative differences in the low excitation energy limit (i.e. Γ μ 0 ). For the strength functions we find distributions going from the Wigner semicircle through gaussians to Breit-Wigner shapes. (orig.)

  10. Rotation in a gravitational billiard

    Science.gov (United States)

    Peraza-Mues, G. G.; Carvente, Osvaldo; Moukarzel, Cristian F.

    Gravitational billiards composed of a viscoelastic frictional disk bouncing on a vibrating wedge have been studied previously, but only from the point of view of their translational behavior. In this work, the average rotational velocity of the disk is studied under various circumstances. First, an experimental realization is briefly presented, which shows sustained rotation when the wedge is tilted. Next, this phenomenon is scrutinized in close detail using a precise numerical implementation of frictional forces. We show that the bouncing disk acquires a spontaneous rotational velocity whenever the wedge angle is not bisected by the direction of gravity. Our molecular dynamics (MD) results are well reproduced by event-driven (ED) simulations. When the wedge aperture angle θW>π/2, the average tangential velocity Rω¯ of the disk scales with the typical wedge vibration velocity vb, and is in general a nonmonotonic function of the overall tilt angle θT of the wedge. The present work focuses on wedges with θW=2π/3, which are relevant for the problem of spontaneous rotation in vibrated disk packings. This study makes part of the PhD Thesis of G. G. Peraza-Mues.

  11. Visual and Haptic Mental Rotation

    Directory of Open Access Journals (Sweden)

    Satoshi Shioiri

    2011-10-01

    Full Text Available It is well known that visual information can be retained in several types of memory systems. Haptic information can also be retained in a memory because we can repeat a hand movement. There may be a common memory system for vision and action. On the one hand, it may be convenient to have a common system for acting with visual information. On the other hand, different modalities may have their own memory and use retained information without transforming specific to the modality. We compared memory properties of visual and haptic information. There is a phenomenon known as mental rotation, which is possibly unique to visual representation. The mental rotation is a phenomenon where reaction time increases with the angle of visual target (eg,, a letter to identify. The phenomenon is explained by the difference in time to rotate the representation of the target in the visual sytem. In this study, we compared the effect of stimulus angle on visual and haptic shape identification (two-line shapes were used. We found that a typical effect of mental rotation for the visual stimulus. However, no such effect was found for the haptic stimulus. This difference cannot be explained by the modality differences in response because similar difference was found even when haptical response was used for visual representation and visual response was used for haptic representation. These results indicate that there are independent systems for visual and haptic representations.

  12. Ultrasonography of the Rotator Cuff

    International Nuclear Information System (INIS)

    Yoon, Yong Cheol

    2006-01-01

    The ultrasonography (US) is an important modality in evaluating shoulder disease. It is accurate in diagnosing the various shoulder diseases including tendinosis, calcific tendinitis, and subacromial bursitis as well as rotator cuff tears. This article presents a pictorial review of US anatomy of the shoulder, the technical aspects of shoulder US, major types of shoulder pathology, and interventional procedure under US guidance

  13. Ultrasonography of the Rotator Cuff

    Energy Technology Data Exchange (ETDEWEB)

    Yoon, Yong Cheol [Samsung Medica Center, Sungkyunkwan University College of Medicine, Seoul (Korea, Republic of)

    2006-09-15

    The ultrasonography (US) is an important modality in evaluating shoulder disease. It is accurate in diagnosing the various shoulder diseases including tendinosis, calcific tendinitis, and subacromial bursitis as well as rotator cuff tears. This article presents a pictorial review of US anatomy of the shoulder, the technical aspects of shoulder US, major types of shoulder pathology, and interventional procedure under US guidance

  14. Rotations in a Vertebrate Setting

    Science.gov (United States)

    McCollum, Gin

    2003-05-01

    Rotational movements of the head are often considered to be measured in a single three dimensional coordinate system implemented by the semicircular canals of the vestibular system of the inner ear. However, the vertebrate body -- including the nervous system -- obeys rectangular symmetries alien to rotation groups. At best, nervous systems mimic the physical rotation group in a fragmented way, only partially reintegrating physical movements in whole organism responses. The vestibular canal reference frame is widely used in nervous systems, for example by eye movements. It is used to some extent even in the cerebrum, as evidenced by the remission of hemineglect -- in which half of space is ignored -- when the vestibular system is stimulated. However, reintegration of space by the organism remains incomplete. For example, compensatory eye movements (which in most cases aid visual fixation) may disagree with conscious self-motion perception. In addition, movement-induced nausea, illusions, and cue-free perceptions demonstrate symmetry breaking or incomplete spatial symmetries. As part of a long-term project to investigate rotation groups in nervous systems, we have analyzed the symmetry group of a primary vestibulo-spinal projection.

  15. Synchrotron Radiation and Faraday Rotation

    NARCIS (Netherlands)

    Heald, George

    2015-01-01

    Synchrotron radiation and its degree of linear polarization are powerful tracers of magnetic fields that are illuminated by cosmic ray electrons. Faraday rotation of the linearly polarized radiation is induced by intervening line-of-sight magnetic fields that are embedded in ionized plasmas. For

  16. Black hole vacua and rotation

    International Nuclear Information System (INIS)

    Krishnan, Chethan

    2011-01-01

    Recent developments suggest that the near-region of rotating black holes behaves like a CFT. To understand this better, I propose to study quantum fields in this region. An instructive approach for this might be to put a large black hole in AdS and to think of the entire geometry as a toy model for the 'near-region'. Quantum field theory on rotating black holes in AdS can be well-defined (unlike in flat space), if fields are quantized in the co-rotating-with-the-horizon frame. First, some generalities of constructing Hartle-Hawking Green functions in this approach are discussed. Then as a specific example where the details are easy to handle, I turn to 2+1 dimensions (BTZ), write down the Green functions explicitly starting with the co-rotating frame, and observe some structural similarities they have with the Kerr-CFT scattering amplitudes. Finally, in BTZ, there is also an alternate construction for the Green functions: we can start from the covering AdS 3 space and use the method of images. Using a 19th century integral formula, I show the equality between the boundary correlators arising via the two constructions.

  17. Perturbative treatment of nuclear rotations

    International Nuclear Information System (INIS)

    Civitarese, O.

    1980-01-01

    In this work, it is described the case corresponding to perturbative quantum treatment of a fermion system in free rotation and the divergences which resulted from the 'break' in symmetry, associated by the adoption of a deformed basis as a non pertubed solution. (A.C.A.S.) [pt

  18. Meniscus Stability in Rotating Systems

    Science.gov (United States)

    Reichel, Yvonne; Dreyer, Michael

    2013-11-01

    In this study, the stability of free surfaces of fluid between two rotating coaxial, circular disks is examined. Radially mounted baffles are used to form menisci of equal size. To the center of the upper disk, a tube is connected in which a separate meniscus is formed. Assuming solid-body rotation and ignoring dynamic effects, it is observed that the free surfaces between the disks fail to remain stable once the rotation speed exceeds a critical value. In other words, Rayleigh-Taylor instability ensues when the capillary forces fail to balance centrifugal forces. Dimensionless critical rotation speeds are studied by means of the Surface Evolver via SE-FIT for varied number of baffles, the normalized distance between the disks, and the normalized central tube radius. Drop tower tests are performed to confirm some of the numerical results. The computation also reveals that there are different modes of instability as a function of the relevant parameters. This study was funded by the space agency of the German Aerospace Center with resources of the Federal Ministry of Economics and Technology on the basis of a resolution of the German Bundestag under grant number 50 RL 1320.

  19. Evaluation using Monte Carlo simulations, of the effect of a shielding, called external shielding, for fotoneutrons generated in linear accelerators, using the computational model of Varian accelerator 2300 C/D operating in eight rotation angles of the GA

    International Nuclear Information System (INIS)

    Silva, Hugo R.; Silva, Ademir X.; Rebello, Wilson F.; Silva, Maria G.

    2011-01-01

    This paper aims to present the results obtained by Monte Carlo simulation of the effect of shielding against neutrons, called External Shielding, to be placed on the heads of linear accelerators used in radiotherapy. For this, it was used the radiation transport code Monte Carlo N-Particle - MCNPX, in which were developed computational model of the head of the linear accelerator Varian 2300 C/D. The equipment was simulated within a bunker, operating at energies of 10, 15 and 18 MV, considering the rotation of the gantry at eight different angles ( 0 deg, 45 deg, 90 deg, 135 deg, 180 deg, 225 deg, 270 deg and 315 deg), in all cases, the equipment was modeled without and with the shielding positioned attached to the head of the accelerator on its bottom. In each of these settings, it was calculated the Ambient Dose Equivalent due to neutron H * (10)n on points situated in the region of the patient (region of interest for evaluation of undesirable neutron doses on the patient) and in the maze of radiotherapy room (region of interest for shielding the access door to the bunker). It was observed for all energies of equipment operation as well as for all angles of inclination of the gantry, a significant reduction in the values of H * (10) n when the equipment operated with the external shielding, both in the region of the patient as in the region of the maze. (author)

  20. Device for measuring atmospheric radon activity

    International Nuclear Information System (INIS)

    Deml, F.; Jansky, Z.; Smejkal, Z.

    1989-01-01

    The device consists of a lightproof case pivoted on a stand. Minimally two chambers are provided in the case for holding glass scintillation chambers, each with an opening in its bottom. The centers of the openings lie on a common circle with its center in the center of rotation of the case. An opening for the passage of light is provided in the stand. The opening also is on the common circle of the centres of the openings. A photomultiplier is placed below the opening. Measurement always proceeds with one scintillation chamber only. Thus, replacement of scintillation chambers can take place simultaneously with measurement and rapid and continuous measurement can be secured. Case rotation can be controlled automatically. The device can be used not only in monitoring air pollution but also in monitoring solid materials, aquifers, mine corridors, natural materials, etc. (J.B.). 1 fig