WorldWideScience

Sample records for photon dose equivalent

  1. The radiobiology of boron neutron capture therapy: Are ''photon-equivalent'' doses really photon-equivalent?

    International Nuclear Information System (INIS)

    Coderre, J.A.; Diaz, A.Z.; Ma, R.

    2001-01-01

    Boron neutron capture therapy (BNCT) produces a mixture of radiation dose components. The high-linear energy transfer (LET) particles are more damaging in tissue than equal doses of low-LET radiation. Each of the high-LET components can multiplied by an experimentally determined factor to adjust for the increased biological effectiveness and the resulting sum expressed in photon-equivalent units (Gy-Eq). BNCT doses in photon-equivalent units are based on a number of assumptions. It may be possible to test the validity of these assumptions and the accuracy of the calculated BNCT doses by 1) comparing the effects of BNCT in other animal or biological models where the effects of photon radiation are known, or 2) if there are endpoints reached in the BNCT dose escalation clinical trials that can be related to the known response to photons of the tissue in question. The calculated Gy-Eq BNCT doses delivered to dogs and to humans with BPA and the epithermal neutron beam of the Brookhaven Medical Research Reactor were compared to expected responses to photon irradiation. The data indicate that Gy-Eq doses in brain may be underestimated. Doses to skin are consistent with the expected response to photons. Gy-Eq doses to tumor are significantly overestimated. A model system of cells in culture irradiated at various depths in a lucite phantom using the epithermal beam is under development. Preliminary data indicate that this approach can be used to detect differences in the relative biological effectiveness of the beam. The rat 9L gliosarcoma cell survival data was converted to photon-equivalent doses using the same factors assumed in the clinical studies. The results superimposed on the survival curve derived from irradiation with Cs-137 photons indicating the potential utility of this model system. (author)

  2. On the calibration of photon dosemeters in the equivalent dose units

    International Nuclear Information System (INIS)

    Bregadze, Yu.I.; Isaev, B.M.; Maslyaev, P.F.

    1980-01-01

    General aspects of transition from exposure dose of photo radiation to equivalent one are considered. By determination the equivalent dose is a function of point location in an irradiated object, that is why it is necessary to know equivalent dose distribution in the human body for uniform description of the risk degree. The international electrotechnical comission recommends to measure equivalent doses at 7 and 800 mg/cm 2 depths in a tissue-equivalent ball with 30 cm diameter, calling them skin equivalent dose and depth equivalent dose, respectively, and to compare them with the permissible 500 mZ and 50 mZ a year, respectively. Practical transition to using equivalent dose for evaluation of radiation danger of being in photon radiation field of low energy should include measures on regraduating already produced dose meters, graduating the dose meters under production and developing the system of their metrologic supply [ru

  3. Investigation of 1-cm dose equivalent for photons behind shielding materials

    International Nuclear Information System (INIS)

    Hirayama, Hideo; Tanaka, Shun-ichi

    1991-03-01

    The ambient dose equivalent at 1-cm depth, assumed equivalent to the 1-cm dose equivalent in practical dose estimations behind shielding slabs of water, concrete, iron or lead for normally incident photons having various energies was calculated by using conversion factors for a slab phantom. It was compared with the 1-cm depth dose calculated with the Monte Carlo code EGS4. It was concluded from this comparison that the ambient dose equivalent calculated by using the conversion factors for the ICRU sphere could be used for the evaluation of the 1-cm dose equivalent for the sphere phantom within 20% errors. Average and practical conversion factors are defined as the conversion factors from exposure to ambient dose equivalent in a finite slab or an infinite one, respectively. They were calculated with EGS4 and the discrete ordinates code PALLAS. The exposure calculated with simple estimation procedures such as point kernel methods can be easily converted to ambient dose equivalent by using these conversion factors. The maximum value between 1 and 30 mfp can be adopted as the conversion factor which depends only on material and incident photon energy. This gives the ambient dose equivalent on the safe side. 13 refs., 7 figs., 2 tabs

  4. Experiences of calibration in photon beams for the personal dose equivalent

    International Nuclear Information System (INIS)

    Lindborg, L.; Gullberg, O.

    1994-01-01

    The calibration quantity shall, according to ICRU, be the personal dose equivalent, H p (d), in a phantom having the composition of the ICRU tissue and the same shape and size as the recommended PMMA calibration phantom, 30 x 30 x 15 cm 3 . There exist differences in backscattering between PMMA and tissue that for certain photon energies could be of importance. This could either be treated as a systematic uncertainty or be incorporated in the definition. However, monoenergetic beams seldom appear in reality and the difference in backscatter is not thought to be important. The calibration quantity for photons was chosen as the absorbed dose to ICRU tissue (times a quality factor 1) at 10 mm depth in a tissue-equivalent slab phantom. In Sweden 13 different services run personal dosimetry. Their initial hesitation about the change of quantity disappeared after testing their photon energy responses. It was found that most TLD systems were measuring the new quantity better than the old one and that the film systems needed only minor corrections. Most TLD systems now report 5% larger dose equivalents for the same irradiation in a photon beam from a 137 Cs source. (author)

  5. Regional inter-comparison of measurements of personal dose equivalent Hp(10) using photon beams

    International Nuclear Information System (INIS)

    Bero, M.; Zahili, M.; Kharita, M.H.

    2012-11-01

    The overall objective is to verify performance and to improve the Individual Monitoring services (IMS). This can be achieved with the following specific objectives of the intercomparison:1. To assess the capabilities of the dosimetry services to measure the quantity H p (10) in photon (gamma and x-ray) fields. 2. To help the participating Member States in achieving sufficiently accurate dosimetry service and, if necessary, 3. To provide guidelines for improvements and not simply a test of the performance of the existing dosimetric service. Actually a significant improvement has been achieved by the participants in the accuracy of evaluating personal dose equivalent from 15% in the first phase to 5% in the second phase. Some participants used the results of the inter-comparison to verify the calibration and to improve their dosimetric procedures, but from the results it was clear that some participants need to a technical support especially in calibration and using their measuring system in the field of personal monitoring. The conclusion contains advises, solutions, propositions and evaluation for all situations which noticed during the intercomparison. (authors)

  6. Measurement of californium-252 gamma photons depth dose distribution in tissue equivalent material. Vol. 4

    Energy Technology Data Exchange (ETDEWEB)

    Fadel, M A; El-Fiki, M A; Eissa, H M; Abdel-Hafez, A; Naguib, S H [National Institute of Standards, Cairo (Egypt)

    1996-03-01

    Phantom of tissue equivalent material with and without bone was used measuring depth dose distribution of gamma-rays from californium-252 source. The source was positioned at center of perspex walled phantom. Depth dose measurements were recorded for X, Y and Z planes at different distances from source. TLD 700 was used for measuring the dose distribution. Results indicate that implantation of bone in tissue equivalent medium cause changes in the gamma depth dose distribution which varies according to variation in bone geometry. 9 figs.

  7. Fluence to Effective Dose and Effective Dose Equivalent Conversion Coefficients for Photons from 50 KeV to 10 GeV

    International Nuclear Information System (INIS)

    Ferrari, A.; Pelliccioni, M.; Pillon, M.

    1996-07-01

    Effective dose equivalent and effective dose per unit photon fluence have been calculated by the FLUKA code for various geometrical conditions of irradiation of an anthropomorphic phantom placed in a vacuum. Calculations have been performed for monoenergetic photons of energy ranging from 50 keV to 10 GeV. The agreement with the results of other authors, when existing, is generally very satisfactory

  8. DEEP code to calculate dose equivalents in human phantom for external photon exposure by Monte Carlo method

    International Nuclear Information System (INIS)

    Yamaguchi, Yasuhiro

    1991-01-01

    The present report describes a computer code DEEP which calculates the organ dose equivalents and the effective dose equivalent for external photon exposure by the Monte Carlo method. MORSE-CG, Monte Carlo radiation transport code, is incorporated into the DEEP code to simulate photon transport phenomena in and around a human body. The code treats an anthropomorphic phantom represented by mathematical formulae and user has a choice for the phantom sex: male, female and unisex. The phantom can wear personal dosimeters on it and user can specify their location and dimension. This document includes instruction and sample problem for the code as well as the general description of dose calculation, human phantom and computer code. (author)

  9. A calculational method of photon dose equivalent based on the revised technical standards of radiological protection law

    International Nuclear Information System (INIS)

    Tanaka, Shun-ichi; Suzuki, Tomoo

    1991-03-01

    The effective conversion factor for photons from 0.03 to 10 MeV were calculated to convert the absorbed dose in air to the 1 cm, 3 mm, and 70 μm depth dose equivalents behind iron, lead, concrete, and water shields up to 30 mfp thickness. The effective conversion factor changes slightly with thickness of the shields and becomes nearly constant at 5 to 10 mfp. The difference of the effective conversion factor was less than 2% between plane normal and point isotropic geometries. It is suggested that the present method, making the data base of the exposure buildup factors useful, would be very effective as compared to a new evaluation of the dose equivalent buildup factors. 5 refs., 7 figs., 22 tabs

  10. The equidosemeter ED-02 as a device for dose equivalent measurements in mixed neutron and photon radiation fields

    International Nuclear Information System (INIS)

    Abrosimov, A.I.; Alekseev, A.G.; Antipov, V.A.; Golovachik, V.T.

    1985-01-01

    The equidosemeter ED-02 is to be used for simultaneous measurements of the dose equivalent, absorbed dose, and mean quality factor of mixed radiations. The detector is a tissue equivalent spherical low-pressure proportional counter tube the signal of which is simultaneously recorded in two channels - a current channel and a pulse one. The current channel is linear and its response proportional to the absorbed dose. The pulse channel includes a nonlinear pulse amplitude converter the characteristic of which, taking into account the required dependence of the mean quality factor on linear energy transfer, has been chosen in such a way that in final counting the pulse channel response is proportional to the difference between dose equivalent and absorbed dose. On the basis of calculations of event spectra in the sensitive volume of the detector, the energy dependence of the dosemeter sensitivity is analysed for neutron energies up to 20 MeV. The characteristic of the nonlinear converter has been calculated on the basis of the construction parameters of the detector and optimized with respect to a representative sample of neutron spectra beyond the shields of nuclear plants. The heterogeneity of the detector, i.e. the difference between the atomic composition of wall and filling and the composition of soft biological tissue as well as the effect of the conducting coating of the case cathode, has been taken into consideration. Moreover, the test results of the device in mixed neutron-photon fields of 60 Co, 239 Pu-α-Be and 252 Cf radioisotope sources are presented. The main measuring error of dose characteristics is shown to be less than 20% in the dose range 1 x 10 -3 to 4 x 10 -3 Sv/h. (author)

  11. Bone-and-muscle-equivalent solid chemical dose meters for photon and electron doses above one kilorad

    International Nuclear Information System (INIS)

    McLaughlin, W.L.; Rosenstein, M.; Levine, H.

    1975-01-01

    Conventional solid dose meters, such as plastic films, powders, emulsions, glasses, ceramics and gels, have a response to ionizing photons and electrons that varies markedly over a broad spectrum when compared with the absorption characteristics of biological tissues. New radiochromic dyed plastic dose meters have been developed with X- and gamma ray and electron energy absorption cross-sections (calculated) and radiation energy responses (experimental) corresponding approximately to those for human muscle and bone, for a spectrum from a few keV to at least 10 MeV. Three-dimensional solid dose meters useful over the absorbed dose range of 10 3 to 10 6 rad are formed by thermosetting a selected combination of monomers containing the radiochromic dye in solution. Thin-film dose meters for the dose range 10 5 to 10 7 rad are formed by casting on optically flat surfaces strippable layers of special combinations of polymers and dyes in solution. The response of these systems to X- and gamma rays and electrons has been studied over various radiation spectra, dose-rates and temperatures during irradiation. (author)

  12. Calculation of dose equivalents for photon skyshine production; Calculo da dose equivalente para fotons decorrente da producao de skyshine

    Energy Technology Data Exchange (ETDEWEB)

    Frota, Marco A.; Kelecom, Alphonse [Universidade Federal Fluminense, Niteroi, RJ (Brazil). Dept. de Biologia Geral. Lab. de Radiobiologia e Radiometria (LARARA)]. E-mail: egbakel@vm.uff.br

    2005-07-01

    Some radiation facilities are designed with little shielding in the ceiling above the accelerator. A problem may then arise as a result of the radiation scattered by the atmosphere to points at ground level outside the treatment room. Stray radiation of this type is referred to as skyshine, and the National Council on Radiation Protection and Measurements Report No. 51 (NCRP 1977) gives methods for the calculation of skyshine for accelerator facilities. McGinley (1993) has compared skyshine measurements made at an 18 MeV medical accelerator facility with values calculated using the techniques presented in NCRP Report No. 51. Measurements were made of the photon levels outside a treatment room housing a Varian 2100 deg C. The roof above the accelerator was designed for weather protection only and offered little shielding for the primary beam and scattered radiation. The distance from the treatment room floor to the roof was 4.27 m, and the primary walls were constructed of concrete 2.0 m thick.The secondary walls were fabricated of concrete 0.99 m thick. The results for the photon skyshine rate dose as a function of distance from the isocenter using Monte Carlo code, are compared with those in NCRP publication 74 and measured obtained. The photon skyshine dose rates simulated for real clinic spectra transmitted through roof range from 4.7 to 14.6 nSv.s{sup -1}. (author)

  13. Measuring instruments of the Physikalisch-Technische Bundesanstalt for realization of the units of the dosimetric quantities standard ion dose, photon-equivalent dose and air-kerma

    International Nuclear Information System (INIS)

    Engelke, B.A.; Oetzmann, W.; Struppek, G.

    1988-08-01

    The realization of the units of the dosimetric quantities exposure, air-kerma and photon-equivalent dose is an important task of the Physikalisch-Technische Bundesanstalt. The report describes the measuring instruments and other technical equipment as well as the determination of the numerous corrections needed. All data and correction factors required for the realization of the units mentioned above are given in many diagrams and tables. (orig.) [de

  14. Effective dose equivalent

    International Nuclear Information System (INIS)

    Huyskens, C.J.; Passchier, W.F.

    1988-01-01

    The effective dose equivalent is a quantity which is used in the daily practice of radiation protection as well as in the radiation hygienic rules as measure for the health risks. In this contribution it is worked out upon which assumptions this quantity is based and in which cases the effective dose equivalent can be used more or less well. (H.W.)

  15. Intercomparison of measurements of personal dose equivalent Hp(10) in photon fields in the West Asia Region

    International Nuclear Information System (INIS)

    2007-01-01

    In accordance with its statutory function, the International Atomic Energy Agency (IAEA) has been assisting its Member States in establishing and upgrading their radiation protection infrastructures, including activities in occupation radiation protection. Individual external dosimetry services for photon radiation have been under establishment or upgrading with support through the Technical Cooperation Model Projects RAW/9/006, Upgrading Radiation Protection Infrastructure (concluded in 2000), and RAW/9/008, Development of Technical Capabilities for Sustainable Radiation and Waste Safety Infrastructure (2001-2004), in all the participating countries in the West Asia Region. Two regional training courses were organized by the IAEA, in Germany in 1998, on Design, Implementation and Management of Individual Monitoring Services (IMS), and in the Syrian Arabic Republic in 2001, on Assessment of Occupational Exposure due to External Sources, under the above stated projects. However, no performance testing has yet been carried out and no regional intercomparisons have been established before in this region. Only two Member States from the region (the Syrian Arab Republic and Lebanon) participated in the interregional Intercomparison for Individual Monitoring of Radiological Measurements for Purposes of Monitoring Personal Dose Equivalent Hp(10) in 1999

  16. Determination of dose equivalent with tissue-equivalent proportional counters

    International Nuclear Information System (INIS)

    Dietze, G.; Schuhmacher, H.; Menzel, H.G.

    1989-01-01

    Low pressure tissue-equivalent proportional counters (TEPC) are instruments based on the cavity chamber principle and provide spectral information on the energy loss of single charged particles crossing the cavity. Hence such detectors measure absorbed dose or kerma and are able to provide estimates on radiation quality. During recent years TEPC based instruments have been developed for radiation protection applications in photon and neutron fields. This was mainly based on the expectation that the energy dependence of their dose equivalent response is smaller than that of other instruments in use. Recently, such instruments have been investigated by intercomparison measurements in various neutron and photon fields. Although their principles of measurements are more closely related to the definition of dose equivalent quantities than those of other existing dosemeters, there are distinct differences and limitations with respect to the irradiation geometry and the determination of the quality factor. The application of such instruments for measuring ambient dose equivalent is discussed. (author)

  17. Determination of the neutron and photon dose equivalent at work places in nuclear facilities of Sweden. An SSI - EURADOS comparison exercise. Part 2: Evaluation

    Energy Technology Data Exchange (ETDEWEB)

    Bartlett, D. [National Radiological Protection Board, Chilton (United Kingdom); Drake, P. [Vattenfall AB, Vaeroebacka (Sweden); Lindborg, L. [Swedish Radiation Protection Inst., Stockholm (Sweden); Klein, H. [Physikalisch-Technische Bundesanstalt, Braunschweig (Germany); Schmitz, Th. [Forschungszentrum Juelich GmbH, Juelich (Germany); Tichy, M

    1999-06-01

    Various mixed neutron-photon fields at workplaces in the containment of pressurised water reactors and in the vicinity of transport containers with spent fuel elements were investigated with spectrometers and dosimeters. The spectral neutron fluences evaluated from measurements with multisphere systems were recommended to be used for the calculation of dosimetric reference values for comparison with the readings of the dosemeters applied simultaneously. It turned out that most of the moderator based area dosemeters overestimated, while the TEPC systems generally underestimated the ambient dose equivalent (DE) values of the rather soft neutron fields encountered at these workplaces. The discrepancies can, however, be explained on the basis of energy dependent responses of the instruments used. The ambient DE values obtained with recently developed area dosemeters based on superheated drop detectors and with track etch based personal dosemeters on phantoms, however, were in satisfying agreement with the reference data. Sets of personal dosemeters simultaneously irradiated on a phantom allowed to roughly estimate the directional dependence of the neutron fluence. Hence, personal and limiting dose equivalent quantities could also be calculated. The personal and ambient DE values were always conservative estimates of the limiting quantities. Unexpectedly, discrepancies were observed for photon DE data measured with GM counters and TEPC systems. The up to 50 % higher readings of the GM counters may be explained by a considerable contribution of high energy photons to the total photon dose equivalent, but photon spectrometry is necessary for final clarification.

  18. Determination of the neutron and photon dose equivalent at work places in nuclear facilities of Sweden. An SSI - EURADOS comparison exercise. Part 2: Evaluation

    International Nuclear Information System (INIS)

    Bartlett, D.; Drake, P.; Lindborg, L.; Klein, H.; Schmitz, Th.; Tichy, M.

    1999-06-01

    Various mixed neutron-photon fields at workplaces in the containment of pressurised water reactors and in the vicinity of transport containers with spent fuel elements were investigated with spectrometers and dosimeters. The spectral neutron fluences evaluated from measurements with multisphere systems were recommended to be used for the calculation of dosimetric reference values for comparison with the readings of the dosemeters applied simultaneously. It turned out that most of the moderator based area dosemeters overestimated, while the TEPC systems generally underestimated the ambient dose equivalent (DE) values of the rather soft neutron fields encountered at these workplaces. The discrepancies can, however, be explained on the basis of energy dependent responses of the instruments used. The ambient DE values obtained with recently developed area dosemeters based on superheated drop detectors and with track etch based personal dosemeters on phantoms, however, were in satisfying agreement with the reference data. Sets of personal dosemeters simultaneously irradiated on a phantom allowed to roughly estimate the directional dependence of the neutron fluence. Hence, personal and limiting dose equivalent quantities could also be calculated. The personal and ambient DE values were always conservative estimates of the limiting quantities. Unexpectedly, discrepancies were observed for photon DE data measured with GM counters and TEPC systems. The up to 50 % higher readings of the GM counters may be explained by a considerable contribution of high energy photons to the total photon dose equivalent, but photon spectrometry is necessary for final clarification

  19. New recommendations for dose equivalent

    International Nuclear Information System (INIS)

    Bengtsson, G.

    1985-01-01

    In its report 39, the International Commission on Radiation Units and Measurements (ICRU), has defined four new quantities for the determination of dose equivalents from external sources: the ambient dose equivalent, the directional dose equivalent, the individual dose equivalent, penetrating and the individual dose equivalent, superficial. The rationale behind these concepts and their practical application are discussed. Reference is made to numerical values of these quantities which will be the subject of a coming publication from the International Commission on Radiological Protection, ICRP. (Author)

  20. Experimental determination of the angular dependence factor for the dose equivalent for photons in calibration phantoms of PMMA

    International Nuclear Information System (INIS)

    Lund, E.; Carlsson, C.A.; Pernicka, F.

    1994-01-01

    The conversion coefficients from air kerma to dose equivalent at a depth of 10 mm in both a spherical and a slab phantom of PMMA have been determined for the X ray qualities: 40, 80 and 295 kV, ISO 'narrow' spectra; and for 137 Cs γ rays. The angular dependence factors have been experimentally determined for the same qualities and for different angles between 0 o and 180 o . The absorbed doses have been measured with thermoluminescence LiF dosemeters. The conversion coefficients and the angular dependence factors are generally found to agree well with calculated ones. Some minor discrepancies are found for the angular dependence factors and the 30 x 30 x 15 cm 3 PMMA slab phantom. (Author)

  1. Open-air ionisation chambers with walls of soft-tissue equivalent material for measuring photon doses

    International Nuclear Information System (INIS)

    Vialettes, H.; Anceau, J.C.; Grand, M.; Petit, G.

    1968-01-01

    The ionisation chambers presented in this report constitute a contribution to research into methods of carrying out correct determinations in the field of health physics. The use of a mixture of teflon containing 42.5 per cent by weight of carbon for the chamber walls makes it possible to measure directly the dose absorbed in air through 300 mg/cm 2 of soft tissue and, consequently, the dose absorbed in the soft tissues with a maximum error of 10 per cent for photon energies of between 10 keV and 10 MeV. Furthermore since this material does not contain hydrogen, the chamber has a sensitivity to neutrons which is much less than other chambers in current use. Finally the shape of these chambers has been studied with a view to obtaining a satisfactory measurement from the isotropy point of view; for example for gamma radiation of 27 keV, the 3 litre chamber is isotropic to within 10 per cent over 270 degrees, and the 12 litre chamber is isotropic to within 10 per cent over 300 degrees; for 1.25 MeV gamma radiation this range is extended over 330 degrees for the 3 litre chamber, and 360 degrees for the 12 litre chamber. This report presents the measurements carried out with these chambers as well as the results obtained. These results are then compared to those obtained with other chambers currently used in the field of health physics. (authors) [fr

  2. What is correct: equivalent dose or dose equivalent

    International Nuclear Information System (INIS)

    Franic, Z.

    1994-01-01

    In Croatian language some physical quantities in radiation protection dosimetry have not precise names. Consequently, in practice either terms in English or mathematical formulas are used. The situation is even worse since the Croatian language only a limited number of textbooks, reference books and other papers are available. This paper compares the concept of ''dose equivalent'' as outlined in International Commission on Radiological Protection (ICRP) recommendations No. 26 and newest, conceptually different concept of ''equivalent dose'' which is introduced in ICRP 60. It was found out that Croatian terminology is both not uniform and unprecise. For the term ''dose equivalent'' was, under influence of Russian and Serbian languages, often used as term ''equivalent dose'' even from the point of view of ICRP 26 recommendations, which was not justified. Unfortunately, even now, in Croatia the legal unit still ''dose equivalent'' defined as in ICRP 26, but the term used for it is ''equivalent dose''. Therefore, in Croatian legislation a modified set of quantities introduced in ICRP 60, should be incorporated as soon as possible

  3. Thermoluminescence dosemeter for personal dose equivalent assessment

    International Nuclear Information System (INIS)

    Silva, T.A. da; Rosa, L.A.R. da; Campos, L.L.

    1995-01-01

    The possibility was investigated of utilising a Brazilian thermoluminescence individual dosemeter, usually calibrated in terms of photon dose equivalent, for the assessment of the personal dose equivalent, H p (d), at depths of 0.07 and 10 mm. The dosemeter uses four CaSO 4 :Dy thermoluminescent detectors, between different filters, as the sensitive materials. It was calibrated in gamma and X radiation fields in the energy range from 17 to 1250 keV. Linear combinations of the responses of three detectors, in this energy range, allow the evaluation of H p (0.07) and H p (10), for radiation incidence angles varying from 0 to 60 degrees, with an accuracy better than 35%. The method is not applicable to mixed photon-beta fields. (author)

  4. Mixed field dose equivalent measuring instruments

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; McDonald, J.C.; Endres, G.W.R.; Quam, W.

    1985-01-01

    In the past, separate instruments have been used to monitor dose equivalent from neutrons and gamma rays. It has been demonstrated that it is now possible to measure simultaneously neutron and gamma dose with a single instrument, the tissue equivalent proportional counter (TEPC). With appropriate algorithms dose equivalent can also be determined from the TEPC. A simple ''pocket rem meter'' for measuring neutron dose equivalent has already been developed. Improved algorithms for determining dose equivalent for mixed fields are presented. (author)

  5. Dose sculpting with generalized equivalent uniform dose

    International Nuclear Information System (INIS)

    Wu Qiuwen; Djajaputra, David; Liu, Helen H.; Dong Lei; Mohan, Radhe; Wu, Yan

    2005-01-01

    With intensity-modulated radiotherapy (IMRT), a variety of user-defined dose distribution can be produced using inverse planning. The generalized equivalent uniform dose (gEUD) has been used in IMRT optimization as an alternative objective function to the conventional dose-volume-based criteria. The purpose of this study was to investigate the effectiveness of gEUD optimization to fine tune the dose distributions of IMRT plans. We analyzed the effect of gEUD-based optimization parameters on plan quality. The objective was to determine whether dose distribution to selected structures could be improved using gEUD optimization without adversely altering the doses delivered to other structures, as in sculpting. We hypothesized that by carefully defining gEUD parameters (EUD 0 and n) based on the current dose distributions, the optimization system could be instructed to search for alternative solutions in the neighborhood, and we could maintain the dose distributions for structures already satisfactory and improve dose for structures that need enhancement. We started with an already acceptable IMRT plan optimized with any objective function. The dose distribution was analyzed first. For structures that dose should not be changed, a higher value of n was used and EUD 0 was set slightly higher/lower than the EUD value at the current dose distribution for critical structures/targets. For structures that needed improvement in dose, a higher to medium value of n was used, and EUD 0 was set to the EUD value or slightly lower/higher for the critical structure/target at the current dose distribution. We evaluated this method in one clinical case each of head and neck, lung and prostate cancer. Dose volume histograms, isodose distributions, and relevant tolerance doses for critical structures were used for the assessment. We found that by adjusting gEUD optimization parameters, the dose distribution could be improved with only a few iterations. A larger value of n could lead to

  6. Calculation methods for determining dose equivalent

    International Nuclear Information System (INIS)

    Endres, G.W.R.; Tanner, J.E.; Scherpelz, R.I.; Hadlock, D.E.

    1987-11-01

    A series of calculations of neutron fluence as a function of energy in an anthropomorphic phantom was performed to develop a system for determining effective dose equivalent for external radiation sources. Critical organ dose equivalents are calculated and effective dose equivalents are determined using ICRP-26 [1] methods. Quality factors based on both present definitions and ICRP-40 definitions are used in the analysis. The results of these calculations are presented and discussed. The effective dose equivalent determined using ICRP-26 methods is significantly smaller than the dose equivalent determined by traditional methods. No existing personnel dosimeter or health physics instrument can determine effective dose equivalent. At the present time, the conversion of dosimeter response to dose equivalent is based on calculations for maximal or ''cap'' values using homogeneous spherical or cylindrical phantoms. The evaluated dose equivalent is, therefore, a poor approximation of the effective dose equivalent as defined by ICRP Publication 26. 3 refs., 2 figs., 1 tab

  7. Editorial: New operational dose equivalent quantities

    International Nuclear Information System (INIS)

    Harvey, J.R.

    1985-01-01

    The ICRU Report 39 entitled ''Determination of Dose Equivalents Resulting from External Radiation Sources'' is briefly discussed. Four new operational dose equivalent quantities have been recommended in ICRU 39. The 'ambient dose equivalent' and the 'directional dose equivalent' are applicable to environmental monitoring and the 'individual dose equivalent, penetrating' and the 'individual dose equivalent, superficial' are applicable to individual monitoring. The quantities should meet the needs of day-to-day operational practice, while being acceptable to those concerned with metrological precision, and at the same time be used to give effective control consistent with current perceptions of the risks associated with exposure to ionizing radiations. (U.K.)

  8. Modification of equivalent photon approximation (EPA) for resolved photon processes

    International Nuclear Information System (INIS)

    Drees, M.; Godbole, R.M.

    1995-05-01

    The authors propose a modification of the equivalent photon approximation (EPA) for processes which involve the parton content of the photon, to take into account the suppression of the photonic parton fluxes due to the virtuality of the photon. They present simple, physically motivated ansaetze to model this suppression and show that even though the parton content of the electron no longer factorizes into an electron flux function and photon structure function, it is still possible to express it as a single integral. They also show that for the TRISTAN (transposable ring intersecting storage accelerators in Nippon) experiments its effects can be numerically of the same size as that of the NLO corrections. Further, it is discussed a possible measurements at HERA (hadron electron ring an large), which can be provide an experimental handle on the effect the authors model through their ansaetze

  9. The definition of the individual dose equivalent

    International Nuclear Information System (INIS)

    Ehrlich, Margarete

    1986-01-01

    A brief note examines the choice of the present definition of the individual dose equivalent, the new operational dosimetry quantity for external exposure. The consequences of the use of the individual dose equivalent and the danger facing the individual dose equivalent, as currently defined, are briefly discussed. (UK)

  10. Calculation methods for determining dose equivalent

    International Nuclear Information System (INIS)

    Endres, G.W.R.; Tanner, J.E.; Scherpelz, R.I.; Hadlock, D.E.

    1988-01-01

    A series of calculations of neutron fluence as a function of energy in an anthropomorphic phantom was performed to develop a system for determining effective dose equivalent for external radiation sources. critical organ dose equivalents are calculated and effective dose equivalents are determined using ICRP-26 methods. Quality factors based on both present definitions and ICRP-40 definitions are used in the analysis. The results of these calculations are presented and discussed

  11. On uncertainties in definition of dose equivalent

    International Nuclear Information System (INIS)

    Oda, Keiji

    1995-01-01

    The author has entertained always the doubt that in a neutron field, if the measured value of the absorbed dose with a tissue equivalent ionization chamber is 1.02±0.01 mGy, may the dose equivalent be taken as 10.2±0.1 mSv. Should it be 10.2 or 11, but the author considers it is 10 or 20. Even if effort is exerted for the precision measurement of absorbed dose, if the coefficient being multiplied to it is not precise, it is meaningless. [Absorbed dose] x [Radiation quality fctor] = [Dose equivalent] seems peculiar. How accurately can dose equivalent be evaluated ? The descriptions related to uncertainties in the publications of ICRU and ICRP are introduced, which are related to radiation quality factor, the accuracy of measuring dose equivalent and so on. Dose equivalent shows the criterion for the degree of risk, or it is considered only as a controlling quantity. The description in the ICRU report 1973 related to dose equivalent and its unit is cited. It was concluded that dose equivalent can be considered only as the absorbed dose being multiplied by a dimensionless factor. The author presented the questions. (K.I.)

  12. Skin Dose Equivalent Measurement from Neutron-Deficient Isotopes

    International Nuclear Information System (INIS)

    Hsu, Hsiao-Hua; Costigan, Steve A.; Romero, Leonard L.; Whicker, Jeffrey J.

    1997-12-01

    Neutron-deficient-isotopes decay via positron emission and/or electron capture often followed by x-ray, gamma-ray, and 0.511 MeV photons from positron annihilation. For cases of significant area and/or personnel contamination with these isotopes, determination of skin dose equivalent (SDE) is required by 10CFR835. For assessment of SDE, we evaluated the MICROSPEC-2(TM) system manufactured by Bubble Technology Industries of Canada which uses three different probes for dose measurement. We used two probes: (1) the X-probe which measures lower energy (4 - 120 keV) photon energy distributions and determines deep dose equivalent, SDE and dose equivalent to eyes, and (2) the B-probe which measures electron (positron) energy distributions, and determines skin dose equivalent. Also, the measured photon and beta spectra can be used to identify radioactive isotopes in the contaminated area. Measurements with several neutron-deficient sources showed that this system provided reasonably accurate SDE rate measurements when compared with calculated benchmark SDE rates with an average percent difference of 40%. Variations were expected because of differences between the assumed geometries used by MlCROSPEC-2 and the calculations when compared to the measurement conditions

  13. Development of dose equivalent meters based on microdosimetric principles

    International Nuclear Information System (INIS)

    Booz, J.

    1984-01-01

    In this paper, the employment of microdosimetric dose-equivalent meters in radiation protection is described considering the advantages of introducing microdosimetric methods into radiation protection, the technical suitability of such instruments for measuring dose equivalent, and finally technical requirements, constraints and solutions together with some examples of instruments and experimental results. The advantage of microdosimetric methods in radiation protection is illustrated with the evaluation of dose-mean quality factors in radiation fields of unknown composition and with the methods of evaluating neutron- and gamma-dose fractions. - It is shown that there is good correlation between dose-mean lineal energy, anti ysub(anti D), and the ICRP quality factor. - Neutron- and gamma-dose fractions of unknown radiation fields can be evaluated with microdosimetric proportional counters without recurrence to other instruments and methods. The problems of separation are discussed. The technical suitability of microdosimetric instruments for measuring dose equivalent is discussed considering the energy response to neutrons and photons and the sensitivity in terms of dose-equivalent rate. Then, considering technical requirements, constraints, and solutions, the problem of the large dynamic range in LET, the large dynamic range in pulse rate, geometry of sensitive volume and electrodes, evaluation of dose-mean quality factors, calibration methods, and uncertainties are discussed. (orig.)

  14. A neutron dose equivalent meter at CAEP

    International Nuclear Information System (INIS)

    Tian Shihai; Lu Yan; Wang Heyi; Yuan Yonggang; Chen Xu

    2012-01-01

    The measurement of neutron dose equivalent has been a widespread need in industry and research. In this paper, aimed at improving the accuracy of neutron dose equivalent meter: a neutron dose counter is simulated with MCNP5, and the energy response curve is optimized. The results show that the energy response factor is from 0.2 to 1.8 for neutrons in the energy range of 2.53×10 -8 MeV to 10 MeV Compared with other related meters, it turns that the design of this meter is right. (authors)

  15. Measurements of the personal dose equivalent

    International Nuclear Information System (INIS)

    Scarlat, F.; Scarisoreanu, A.; Badita, E.; Oane, M.; Mitru, E.

    2008-01-01

    Full text: The paper presents the results of measurements related to the personal dose equivalent in the rooms adjacent to NILPRP 7 MeV linear accelerator, by means of the secondary standard chamber T34035 Hp(10). The chamber was calibrated by PTB at S- 137 Cs (E av = 661.6 keV, T 1/2 11050 days) and has N H = 3.17x10 6 Sv/C calibration factor for the personal dose equivalent, Hp(10), at a depth of 10 mm in climatic reference conditions. The measurements were made for the two operation mode of the 7 MeV linac: electrons and bremsstrahlung

  16. Comparison of conversion coefficients for equivalent dose in terms of air kerma using a sitting and standing female adult voxel simulators exposure to photons in antero-posterior irradiation geometry

    International Nuclear Information System (INIS)

    Cavalcante, F.R.; Galeano, D.C.; Carvalho Júnior, A.B.; Hunt, J.

    2014-01-01

    Due to the difficulty in implementing invasive techniques for calculations of dose for some exposure scenarios, computational simulators have been created to represent as realistically as possible the structures of the human body and through radiation transport simulations to obtain conversion coefficients (CCs) to estimate dose. In most published papers simulators are implemented in the standing posture and this may not describe a real scenario of exposure. In this work we developed exposure scenarios in the Visual Monte Carlo (VMC) code using a female simulator in standing and sitting postures. The simulator was irradiated in the antero-posterior (AP) geometry by a plane source of monoenergetic photons with energy from 10 keV to 2 MeV. The conversion coefficients for equivalent dose in terms of air kerma (H T /K air ) were calculated for both scenarios and compared. The results show that the percentage difference of CCs for the organs of the head and thorax was not significant (less than 5%) since the anatomic position of the organs is the same in both postures. The percentage difference is more significant to the ovaries (71% for photon energy of 20 keV), to the bladder (39% at 60 keV) and to the uterus (37% at 100 keV) due to different processes of radiation interactions in the legs of the simulator when its posture is changed. For organs and tissues that are distributed throughout the entire body, such as bone (21% at 100 keV) and muscle (30% at 80 keV) the percentage difference of CCs reflects a reduction of interaction of photons with the legs of the simulator. Therefore, the calculation of conversion coefficients using simulators in the sitting posture is relevant for a more accurate dose estimation in real exposures to radiation. - Highlights: ► Scenarios of external photon exposures were performed in VMC code. ► The FAX simulator was irradiated in sitting and standing postures. ► The irradiation geometry used was the antero-posterior (AP). ► The

  17. Comparison of conversion coefficients for equivalent dose in terms of air kerma for photons using a male adult voxel simulator in sitting and standing posture with geometry of irradiation antero-posterior

    International Nuclear Information System (INIS)

    Galeano, D.C.; Cavalcante, F.R.; Carvalho, A.B.; Hunt, J.

    2014-01-01

    The dose conversion coefficient (DCC) is important to quantify and assess effective doses associated with medical, professional and public exposures. The calculation of DCCs using anthropomorphic simulators and radiation transport codes is justified since in-vivo measurement of effective dose is extremely difficult and not practical for occupational dosimetry. DCCs have been published by the ICRP using simulators in a standing posture, which is not always applicable to all exposure scenarios, providing an inaccurate dose estimation. The aim of this work was to calculate DCCs for equivalent dose in terms of air kerma (H/Kair) using the Visual Monte Carlo (VMC) code and the VOXTISS8 adult male voxel simulator in sitting and standing postures. In both postures, the simulator was irradiated by a plane source of monoenergetic photons in antero-posterior (AP) geometry. The photon energy ranged from 15 keV to 2 MeV. The DCCs for both postures were compared and the DCCs for the standing simulator were higher. For certain organs, the difference of DCCs were more significant, as in gonads (48% higher), bladder (16% higher) and colon (11% higher). As these organs are positioned in the abdominal region, the posture of the anthropomorphic simulator modifies the form in which the radiation is transported and how the energy is deposited. It was also noted that the average percentage difference of conversion coefficients was 33% for the bone marrow, 11% for the skin, 13% for the bone surface and 31% for the muscle. For other organs, the percentage difference of the DCCs for both postures was not relevant (less than 5%) due to no anatomical changes in the organs of the head, chest and upper abdomen. We can conclude that is important to obtain DCCs using different postures from those present in the scientific literature. - Highlights: • Scenarios of external photon exposures were performed in VMC code. • The VOXTISS8 simulator was irradiated in standing and sitting postures.

  18. The Use Of Optical Properties Of Cr-39 In Alpha Particle Equivalent Dose Measurements

    International Nuclear Information System (INIS)

    Shnishin, K.A.

    2007-01-01

    In this work, optical properties of alpha irradiated Cr-39 were measured as a function of optical photon wavelength from 200-1100 nm. Optical energy gap and optical absorption at finite wavelength was also calculated and correlated to alpha fluence and dose equivalent. Alpha doses were calculated from the corresponding irradiation fluence and specific energy loss using TRIM computer program. It was found that, the optical absorption of unattached Cr-39 was varied with alpha fluence and corresponding equivalent doses. Also the optical energy gab was varied with fluence and dose equivalent of alpha particles. This work introduces a reasonably simple method for the Rn dose equivalent calculation by Cr-39 track

  19. On the validity of the equivalent-photon approximation for virtual photon-photon collisions

    International Nuclear Information System (INIS)

    Carimalo, C.; Kessler, P.; Parisi, J.

    1979-05-01

    For virtual photon-photon collisions in electron storage rings, one derive the equivalent-photon approximation from a helicity treatment, and present it in two forms, involving respectively (i) polarized transverse photons ('transverse-photon approximation') and (ii) unpolarized ones ('Williams-Weizsaecker approximation'). One first postulates the conditions of validity of the approximation on the basis of analytic considerations, and then check them numerically in the case of the process e e → e e μ + μ - . For this check, we consider the completely differentiated cross section as far as approximation (i) is concerned; and in the case of approximation (ii), the cross section differentiated with respect to all variables except the azimuthal angles. The results are given in the form of Tables showing the lower and higher limit of the error involved in the approximation for a large variety of kinematic configurations (i. e., energy losses and scattering angles of both electrons). Those Tables are discussed in detail, and conclusions are drawn as to the applicability of the equivalent-photon approximation to future experiments to be performed with high-energy electron storage rings

  20. A comparison of the angular dependence of effective dose and effective dose equivalent

    International Nuclear Information System (INIS)

    Sitek, M.A.; Gierga, D.P.; Xu, X.G.

    1996-01-01

    In ICRP (International Commission on Radiological Protection) Publication 60, the set of critical organs and their weighing factors were changed, defining the quantity effective dose, E. This quantity replaced the effective dose equivalent, H E , as defined by ICRP 26. Most notably, the esophagus was added to the list of critical organs. The Monte Carlo neutron/photon transport code MCNP was used to determine the effective dose to sex-specific anthropomorphic phantoms. The phantoms, developed in previous research, were modified to include the esophagus. Monte Carlo simulations were performed for monoenergetic photon beams of energies 0.08 MeV, 0.3 MeV, and 1.0 MeV for various azimuthal and polar angles. Separate organ equivalent doses were determined for male and female phantoms. The resulting organ equivalent doses were calculated from arithmetic mean averages. The angular dependence of effective dose was compared with that of effective dose equivalent reported in previous research. The differences between the two definitions and possible implications to regulatory agencies were summarized

  1. Dose equivalent distribution during occupational exposure in oncology

    International Nuclear Information System (INIS)

    Marco H, J.

    1996-01-01

    In this work are presented the results of the radiological surveillance of occupationally exposed workers at the National Institute of Oncology and Radiology during 26 years. The incidence of the equivalent dose in the personal working with radiant sources and radioactive substances in areas of x rays diagnostic, teletherapy, brachytherapy, nuclear medicine and biomedical research was showed. The employed dosimetric system makes use of ORWO RD3/RD4 monitoring film with copper and lead filters inside a plastic cassette manufactured in Cuba. The experimental method is supported by the optical densitometric analysis of films together with a set of standard film calibrated in standard X and gamma photon beams by means of a secondary standard dosimeter, type NPL. Statistics show that except those workings with radium-226, manual brachytherapy or Mo-99/Tc-99 generator elution, the equivalent dose distribution in our workers has been kept in regions well down the annual permissible limit. (authors). 6 refs., 3 tabs

  2. Open-air ionisation chambers with walls of soft-tissue equivalent material for measuring photon doses; Chambres d'ionisation d'ambiance a parois en materiau equivalent aux tissus mous pour la mesure des doses absorbees dues aux photons

    Energy Technology Data Exchange (ETDEWEB)

    Vialettes, H.; Anceau, J.C.; Grand, M.; Petit, G. [Commissariat a l' Energie Atomique, Saclay (France). Centre d' Etudes Nucleaires

    1968-07-01

    The ionisation chambers presented in this report constitute a contribution to research into methods of carrying out correct determinations in the field of health physics. The use of a mixture of teflon containing 42.5 per cent by weight of carbon for the chamber walls makes it possible to measure directly the dose absorbed in air through 300 mg/cm{sup 2} of soft tissue and, consequently, the dose absorbed in the soft tissues with a maximum error of 10 per cent for photon energies of between 10 keV and 10 MeV. Furthermore since this material does not contain hydrogen, the chamber has a sensitivity to neutrons which is much less than other chambers in current use. Finally the shape of these chambers has been studied with a view to obtaining a satisfactory measurement from the isotropy point of view; for example for gamma radiation of 27 keV, the 3 litre chamber is isotropic to within 10 per cent over 270 degrees, and the 12 litre chamber is isotropic to within 10 per cent over 300 degrees; for 1.25 MeV gamma radiation this range is extended over 330 degrees for the 3 litre chamber, and 360 degrees for the 12 litre chamber. This report presents the measurements carried out with these chambers as well as the results obtained. These results are then compared to those obtained with other chambers currently used in the field of health physics. (authors) [French] Les chambres d'ionisation presentees dans ce rapport apportent une contribution a la recherche de moyens dosimetriques adaptes aux mesures a effectuer pour assurer une dosimetrie correcte dans le domaine de la radioprotection. L'utilisation d'un melange de teflon charge a 42.5 pour cent en masse de carbone comme materiau constituant les parois de la chambre permet de realiser un dosimetre mesurant directement la dose absorbee dans l'air sous 3OO mg/cm{sup 2} de tissu mou et, par consequent, la dose absorbee dans les tissus mous avec une erreur maximale de 10 pour cent, pour des photons d

  3. Neutrons in active proton therapy. Parameterization of dose and dose equivalent

    Energy Technology Data Exchange (ETDEWEB)

    Schneider, Uwe; Haelg, Roger A. [Univ. of Zurich (Switzerland). Dept. of Physics; Radiotherapy Hirslanden AG, Aarau (Switzerland); Lomax, Tony [Paul Scherrer Institute, Villigen (Switzerland). Center for Proton Therapy

    2017-08-01

    One of the essential elements of an epidemiological study to decide if proton therapy may be associated with increased or decreased subsequent malignancies compared to photon therapy is an ability to estimate all doses to non-target tissues, including neutron dose. This work therefore aims to predict for patients using proton pencil beam scanning the spatially localized neutron doses and dose equivalents. The proton pencil beam of Gantry 1 at the Paul Scherrer Institute (PSI) was Monte Carlo simulated using GEANT. Based on the simulated neutron dose and neutron spectra an analytical mechanistic dose model was developed. The pencil beam algorithm used for treatment planning at PSI has been extended using the developed model in order to calculate the neutron component of the delivered dose distribution for each treated patient. The neutron dose was estimated for two patient example cases. The analytical neutron dose model represents the three-dimensional Monte Carlo simulated dose distribution up to 85 cm from the proton pencil beam with a satisfying precision. The root mean square error between Monte Carlo simulation and model is largest for 138 MeV protons and is 19% and 20% for dose and dose equivalent, respectively. The model was successfully integrated into the PSI treatment planning system. In average the neutron dose is increased by 10% or 65% when using 160 MeV or 177 MeV instead of 138 MeV. For the neutron dose equivalent the increase is 8% and 57%. The presented neutron dose calculations allow for estimates of dose that can be used in subsequent epidemiological studies or, should the need arise, to estimate the neutron dose at any point where a subsequent secondary tumour may occur. It was found that the neutron dose to the patient is heavily increased with proton energy.

  4. INTERCOMPARISON ON THE MEASUREMENT OF THE QUANTITY PERSONAL DOSE EQUIVALENT HP(10) IN PHOTON FIELDS. LINEARITY DEPENDENCE, LOWER LIMIT OF DETECTION AND UNCERTAINTY IN MEASUREMENT OF DOSIMETRY SYSTEMS OF INDIVIDUAL MONITORING SERVICES IN GABON AND GHANA.

    Science.gov (United States)

    Ondo Meye, P; Schandorf, C; Amoako, J K; Manteaw, P O; Amoatey, E A; Adjei, D N

    2017-12-01

    An inter-comparison study was conducted to assess the capability of dosimetry systems of individual monitoring services (IMSs) in Gabon and Ghana to measure personal dose equivalent Hp(10) in photon fields. The performance indicators assessed were the lower limit of detection, linearity and uncertainty in measurement. Monthly and quarterly recording levels were proposed with corresponding values of 0.08 and 0.025 mSv, and 0.05 and 0.15 mSv for the TLD and OSL systems, respectively. The linearity dependence of the dosimetry systems was performed following the requirement given in the Standard IEC 62387 of the International Electrotechnical Commission (IEC). The results obtained for the two systems were satisfactory. The procedure followed for the uncertainty assessment is the one given in the IEC technical report TR62461. The maximum relative overall uncertainties, in absolute value, expressed in terms of Hp(10), for the TL dosimetry system Harshaw 6600, are 44. 35% for true doses below 0.40 mSv and 36.33% for true doses ≥0.40 mSv. For the OSL dosimetry system microStar, the maximum relative overall uncertainties, in absolute value, are 52.17% for true doses below 0.40 mSv and 37.43% for true doses ≥0.40 mSv. These results are in good agreement with the requirements for accuracy of the International Commission on Radiological protection. When expressing the uncertainties in terms of response, comparison with the IAEA requirements for overall accuracy showed that the uncertainty results were also acceptable. The values of Hp(10) directly measured by the two dosimetry systems showed a significant underestimation for the Harshaw 6600 system, and a slight overestimation for the microStar system. After correction for linearity of the measured doses, the two dosimetry systems gave better and comparable results. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  5. Determination of the equivalent of environmental dose, H*(d), in a radiotherapy installation

    International Nuclear Information System (INIS)

    Lima, M.A.F.; Borges, J.C.; Mota, H.C.

    1998-01-01

    In order to put into practice radiological protection has been required conversion factors for environmental dose equivalent determination to air kerma value for different kinds of photon and electron beams, such dose values have been determined in a spheric phantom of 30 cm diameter in a alignment field and expanded in a depth of this sphere. Details will be given for determining of equivalent dose distribution calculation using Monte Carlo computational method (ESG4) following the recommendations of ICRU. (Author)

  6. Development of air equivalent gamma dose monitor

    International Nuclear Information System (INIS)

    Alex, Mary; Bhattacharya, Sadhana; Karpagam, R.; Prasad, D.N.; Jakati, R.K.; Mukhopadhyay, P.K.; Patil, R.K.

    2010-01-01

    The paper describes design and development of air equivalent gamma absorbed dose monitor. The monitor has gamma sensitivity of 84 pA/R/h for 60 Co source. The characterization of the monitor has been done to get energy dependence on gamma sensitivity and response to gamma radiation field from 1 R/hr to 5000 R/hr. The gamma sensitivity in the energy range of 0.06 to 1.25MeV relative to 137 Cs nuclide was within 2.5%. The linearity of the monitor response as a function of gamma field from 10 R/h to 3.8 kR/h was within 6%. The monitor has been designed for its application in harsh environment. It has been successfully qualified to meet environmental requirements of shock. (author)

  7. Accuracy limits of the equivalent field method for irregular photon fields

    International Nuclear Information System (INIS)

    Sanz, Dario Esteban

    2002-01-01

    A mathematical approach is developed to evaluate the accuracy of the equivalent field method using basic clinical photon beam data. This paper presents an analytical calculation of dose errors arising when field equivalencies, calculated at a certain reference depth, are translated to other depths. The phantom scatter summation is expressed as a Riemann-Stieltjes integral and two categories of irregular fields are introduced: uniform and multiform. It is shown that multiform fields produce errors whose magnitudes are nearly twice those corresponding to uniform fields in extreme situations. For uniform field shapes, the maximum, local, relative dose errors, when the equivalencies are calculated at 10 cm depth on the central axis and translated to a depth of 30 cm, are 3.8% and 8.8% for 6 MV and cobalt-60 photon beams, respectively. In terms of maximum dose those errors are within 1-2%. This supports the conclusion that the equivalencies between rectangular fields, which are examples of uniform fields, are applicable to dose ratio functions irrespective of beam energy. However, the magnitude of such errors could be of importance when assessing the exit dose for in vivo monitoring. This work provides a better understanding of the influence of the irregular field shapes on the accuracy of the equivalent field method. (author)

  8. Dose equivalent distributions in the AAEC total body nitrogen facility

    International Nuclear Information System (INIS)

    Allen, B.J.; Bailey, G.M.; McGregor, B.J.

    1985-01-01

    The incident neutron dose equivalent in the AAEC total body nitrogen facility is measured by a calibrated remmeter. Dose equivalent rates and distributions are calculated by Monte Carlo techniques which take account of the secondary neutron flux from the collimator. Experiment and calculation are found to be in satisfactory agreement. The effective dose equivalent per exposure is determined by weighting organ doses, and the potential detriment per exposure is calculated from ICRP risk factors

  9. Equivalence principle and quantum mechanics: quantum simulation with entangled photons.

    Science.gov (United States)

    Longhi, S

    2018-01-15

    Einstein's equivalence principle (EP) states the complete physical equivalence of a gravitational field and corresponding inertial field in an accelerated reference frame. However, to what extent the EP remains valid in non-relativistic quantum mechanics is a controversial issue. To avoid violation of the EP, Bargmann's superselection rule forbids a coherent superposition of states with different masses. Here we suggest a quantum simulation of non-relativistic Schrödinger particle dynamics in non-inertial reference frames, which is based on the propagation of polarization-entangled photon pairs in curved and birefringent optical waveguides and Hong-Ou-Mandel quantum interference measurement. The photonic simulator can emulate superposition of mass states, which would lead to violation of the EP.

  10. Characteristics of natural background external radiation and effective dose equivalent

    International Nuclear Information System (INIS)

    Fujimoto, Kenzo

    1989-01-01

    The two sources of natural radiation - cosmic rays and primordial radionuclides - are described. The factors affecting radiation doses received from natural radiation and the calculation of effective dose equivalent due to natural radiation are discussed. 10 figs., 3 tabs

  11. Calculation of committed dose equivalent from intake of tritiated water

    International Nuclear Information System (INIS)

    Law, D.V.

    1978-08-01

    A new computerized method of calculating the committed dose equivalent from the intake of tritiated water at Harwell is described in this report. The computer program has been designed to deal with a variety of intake patterns and urine sampling schemes, as well as to produce committed dose equivalents corresponding to any periods for which individual monitoring for external radiation is undertaken. Details of retrospective doses are added semi-automatically to the Radiation Dose Records and committed dose equivalents are retained on a separate file. (author)

  12. Basis for calculating body equivalent doses after external radiation exposure. 3. rev. and enl. ed.; Berechnungsgrundlage fuer die Ermittlung von Koerper-Aequivalentdosen bei aeusserer Strahlenexposition

    Energy Technology Data Exchange (ETDEWEB)

    Sarenio, O. (comp.) [Geschaeftsstelle der Strahlenschutzkommission beim Bundesamt fuer Strahlenschutz, Bonn (Germany)

    2017-07-01

    The book on the basis for calculating body equivalent doses after external radiation exposure includes the following issues: introduction covering the scope of coverage and body equivalent doses for radiation protection, terminology, photon radiation, neutron radiation, electron radiation, mixed radiation fields and the estimation of body equivalent doses for skin surface contamination.

  13. Mathematical simulation of biologically equivalent doses for LDR-HDR

    International Nuclear Information System (INIS)

    Slosarek, K.; Zajusz, A.

    1996-01-01

    Based on the LQ model examples of biologically equivalent doses LDR, HDR and external beams were calculated. The biologically equivalent doses for LDR were calculated by appending to the LQ model the corrector for the time of repair of radiation sublethal damages. For radiation continuously delivered at a low dose rate the influence of sublethal damage repair time changes on biologically equivalent doses were analysed. For fractionated treatment with high dose rate the biologically equivalent doses were calculated by adding to the LQ model the formula of accelerated repopulation. For total biologically equivalent dose calculation for combine LDR-HDR-Tele irradiation examples are presented with the use of different parameters of the time of repair of sublethal damages and accelerated repopulation. The calculations performed show, that the same biologically equivalent doses can be obtained for different parameters of cell kinetics changes during radiation treatment. It also shows, that during biologically equivalent dose calculations for different radiotherapy schedules, ignorance of cell kinetics parameters can lead to relevant errors

  14. A phantom for assessing the personal dose equivalent, HP(10)

    International Nuclear Information System (INIS)

    Santoro, C.; Filho, J.A

    2013-01-01

    Characteristics of a phantom designed to evaluate the personal dose equivalent, H P (10), and appropriate for photographic dosimetry are presented. It is called HP(10) phantom due to cavities constructed to insert dosimetric films at a depth of 10 mm. The H P (10) phantom is irradiated with ionizing radiation energy, E, from 45 to 1250 keV, with doses ranging from 0.2 to 50 mSv. It is positioned in the direction α = 0 °, and the radiation field focusing perpendicular to its front surface. So, are established calibration curves of dosimeters in the position conventionally true and quantities H P (10). It made a comparison between the responses obtained with the H P (10) phantom and responses obtained when using the calibration procedure recommended by ISO dosimeters. The ISO recommends getting the air kerma, Ka, for photons at test point of the radiation field by an ionization chamber. And through conversion coefficients, h pK (10; E, α), becomes the air kerma for H P (10). The ISO 4037-3 recommendation has been studied by researchers to ensure that the low energy spectral differences occur in radiation fields which are generated by various X-ray equipment, and induce changes in the percentages of conversion coefficients on the order of 10% to 90% . On the basis of the recommendations ISO, this article develops phantom able to assess the dose to the influence of scattering and absorption of radiation, its implications with respect to dosimetry, providing improvement in the assessment of doses. (author)

  15. Equivalent-spherical-shield neutron dose calculations

    International Nuclear Information System (INIS)

    Russell, G.J.; Robinson, H.

    1988-01-01

    Neutron doses through 162-cm-thick spherical shields were calculated to be 1090 and 448 mrem/h for regular and magnetite concrete, respectively. These results bracket the measured data, for reinforced regular concrete, of /approximately/600 mrem/h. The calculated fraction of the high-energy (>20 MeV) dose component also bracketed the experimental data. The measured and calculated doses were for a graphite beam stop bombarded with 100 nA of 800-MeV protons. 6 refs., 2 figs., 1 tab

  16. Annual average equivalent dose of workers form health area

    International Nuclear Information System (INIS)

    Daltro, T.F.L.; Campos, L.L.

    1992-01-01

    The data of personnel monitoring during 1985 and 1991 of personnel that work in health area were studied, obtaining a general overview of the value change of annual average equivalent dose. Two different aspects were presented: the analysis of annual average equivalent dose in the different sectors of a hospital and the comparison of these doses in the same sectors in different hospitals. (C.G.C.)

  17. Simulation of lung cancer treatment with equivalent dose calculation and analysis of the dose distribution profile

    International Nuclear Information System (INIS)

    Thalhofer, J. L.; Marques L, J.; Da Silva, A. X.; Dos Reis J, J. P.; Da Silva J, W. F. R.; Arruda C, S. C.; Monteiro de S, E.; Santos B, D. V.

    2017-10-01

    Actually, lung cancer is one of the most lethal types, due to the disease in the majority of the cases asymptomatic in the early stages, being the detection of the pathology in advanced stage, with tumor considerable volume. Dosimetry analysis of healthy organs under real conditions is not feasible. Therefore, computational simulations are used to auxiliary in dose verification in organs of patients submitted to radiotherapy. The goal of this study is to calculate the equivalent dose, due to photons, in surrounding in healthy organs of a patient submitted to radiotherapy for lung cancer, through computational modeling. The simulation was performed using the MCNPX code (Version, 2006], Rex and Regina phantom [ICRP 110, 2008], radiotherapy room, Siemens Oncor Expression accelerator operating at 6 MV and treatment protocol adopted at the Inca (National Cancer Institute, Brazil). The results obtained, considering the dose due to photons for both phantom indicate that organs located inside the thoracic cavity received higher dose, being the bronchi, heart and esophagus more affected, due to the anatomical positioning. Clinical data describe the development of bronchiolitis, esophagitis, and cardiomyopathies with decreased cardiopulmonary function as one of the major effects of lung cancer treatment. In the Regina phantom, the second largest dose was in the region of the breasts with 615,73 mSv / Gy, while in the Rex 514,06 mSv / Gy, event related to the difference of anatomical structure of the organ. Through the t mesh command, a qualitative analysis was performed between the dose deposition profile of the planning system and the simulated treatment, with a similar profile of the dose distribution being verified along the patients body. (Author)

  18. Committed dose equivalent in the practice of radiological protection

    International Nuclear Information System (INIS)

    Nenot, J.C.; Piechowski, J.

    1985-01-01

    In the case of internal exposure, the dose is not received at the moment of exposure, as happens with external exposure, since the incorporated radionuclide irradiates the various organs and tissues during the time it is present in the body. By definition, the committed dose equivalent corresponds to the received dose integrated over 50 years from the date of intake. In order to calculate it, one has to know the intake activity and the value of the committed dose equivalent per unit of intake activity. The uncertainties of the first parameter are such that the committed dose equivalent can only be regarded as an order of magnitude and not as a very accurate quantity. The use of it is justified, however, for, like the dose equivalent for external exposure, it expresses the risk of stochastic effects for the individual concerned since these effects, should they appear, would do so only after a latent period which is generally longer than the dose integration time. Moreover, the use of the committed dose equivalent offers certain advantages for dosimetric management, especially when it is simplified. A practical problem which may arise is that the annual dose limit is apparently exceeded by virtue of the fact that one is taking account, in the first year, of doses which will actually be received only in the following years. These problems are rare enough in practice to be dealt with individually in each case. (author)

  19. Time-Dependent Neutron and Photon Dose-Field Analysis

    Energy Technology Data Exchange (ETDEWEB)

    Wooten, Hasani Omar [Georgia Inst. of Technology, Atlanta, GA (United States)

    2005-08-01

    A unique tool is developed that allows the user to model physical representations of complicated glovebox facilities in two dimensions and determine neutral-particle flux and ambient dose-equivalent fields throughout that geometry. The Pandemonium code, originally designed to determine flux and dose-rates only, is improved to include realistic glovebox geometries, time-dependent source and detector positions, time-dependent shielding thickness calculations, time-integrated doses, a representative criticality accident scenario based on time-dependent reactor kinetics, and more rigorous photon treatment. A primary benefit of this work has been an extensive analysis and improvement of the photon model that is not limited to the application described in this thesis. The photon model has been extended in energy range to 10 MeV to include photons from fission and new photon buildup factors have been included that account for the effects of photon buildup at slant-path thicknesses as a function of angle, where the mean free path thickness has been preserved. The overall system of codes is user-friendly and it is directly applicable to facilities such as the plutonium facility at Los Alamos National Laboratory, where high-intensity neutron and photon emitters are regularly used. The codes may be used to determine a priori doses for given work scenarios in an effort to supply dose information to process models which will in turn assist decision makers on ensuring as low as reasonably achievable (ALARA) compliance. In addition, coupling the computational results of these tools with the process model visualization tools will help to increase worker safety and radiological safety awareness.

  20. Assessing the effect of electron density in photon dose calculations

    International Nuclear Information System (INIS)

    Seco, J.; Evans, P. M.

    2006-01-01

    Photon dose calculation algorithms (such as the pencil beam and collapsed cone, CC) model the attenuation of a primary photon beam in media other than water, by using pathlength scaling based on the relative mass density of the media to water. In this study, we assess if differences in the electron density between the water and media, with different atomic composition, can influence the accuracy of conventional photon dose calculations algorithms. A comparison is performed between an electron-density scaling method and the standard mass-density scaling method for (i) tissues present in the human body (such as bone, muscle, etc.), and for (ii) water-equivalent plastics, used in radiotherapy dosimetry and quality assurance. We demonstrate that the important material property that should be taken into account by photon dose algorithms is the electron density, and not the mass density. The mass-density scaling method is shown to overestimate, relative to electron-density predictions, the primary photon fluence for tissues in the human body and water-equivalent plastics, where 6%-7% and 10% differences were observed respectively for bone and air. However, in the case of patients, differences are expected to be smaller due to the large complexity of a treatment plan and of the patient anatomy and atomic composition and of the smaller thickness of bone/air that incident photon beams of a treatment plan may have to traverse. Differences have also been observed for conventional dose algorithms, such as CC, where an overestimate of the lung dose occurs, when irradiating lung tumors. The incorrect lung dose can be attributed to the incorrect modeling of the photon beam attenuation through the rib cage (thickness of 2-3 cm in bone upstream of the lung tumor) and through the lung and the oversimplified modeling of electron transport in convolution algorithms. In the present study, the overestimation of the primary photon fluence, using the mass-density scaling method, was shown

  1. The meaning and the principle of determination of the effective dose equivalent in radiation protection

    International Nuclear Information System (INIS)

    Drexler, G.; Williams, G.; Zankl, M.

    1985-01-01

    Since the introduction of the quantity ''effective dose equivalent'' within the framework of new radiation concepts, the meaning and interpretation of the quantity is often discussed and debated. Because of its adoption as a limiting quantity in many international and national laws, it is necessary to be able to interpret this main radiation protection quantity. Examples of organ doses and the related Hsub(E) values in occupational and medical exposures are presented and the meaning of the quantity is considered for whole body exposures to external and internal photon sources, as well as for partial body external exposures to photons. (author)

  2. The cost of collective dose equivalent

    International Nuclear Information System (INIS)

    Clark, M.J.; Fleishman, A.B.

    1979-01-01

    The successful application of the ALARA principle is dependent on a monetary evaluation of collective dose, so that the costs of control may be directly compared with the costs of detriment. The paper initially reviews the traditional, quantitative methods of valuing human life from which a monetary value of the man-sievert can be derived by subsequent application of risk factors. The political, ethical and economic implications of employing such techniques in radiological protection are noted. An alternative approach to the problem is then outlined, based on estimating the marginal value of a change in risk. At low levels of individual risk this includes only the tangible, economic consequences of detriment from output losses and medical costs. However, as risk levels rise, social costs related to the anxiety associated with radiation-induced health effects are also incorporated and the valuation increases disproportionally. While such valuations are in principle measurable, in the absence of empirical data a risk/detriment cost relationship is proposed, leading to a variable monetary value of the man-sievert which is a function of the per caput dose of a given population sub-group. Application of this methodology will help to ensure that radiological protection resources are spent in areas where they are most required. (author)

  3. Effective dose equivalents from external radiation due to Chernobyl accident

    International Nuclear Information System (INIS)

    Erkin, V.G.; Debedev, O.V.; Balonov, M.I.; Parkhomenko, V.I.

    1992-01-01

    Summarized data on measurements of individual dose of external γ-sources in 1987-1990 of population of western areas of Bryansk region were presented. Type of distribution of effective dose equivalent, its significance for various professional and social groups of population depending on the type of the house was discussed. Dependences connecting surface soil activity in the populated locality with average dose of external radiation sources were presented. Tendency of dose variation in 1987-1990 was shown

  4. Dose characteristics of high-energy electrons, muons and photons

    International Nuclear Information System (INIS)

    Britvich, G.I.; Krupnyj, G.I.; Peleshko, V.N.; Rastsvetalov, Ya.N.

    1980-01-01

    Differential distribution of energy release at different depth of tissue-equivalent phantoms (plexiglas, polystyrene, polyethylene) at the energy of incident electrons, muons of 0.2-40 GeV and photons with the mean energy of 3.6 GeV are measured. The error of experimental results does not exceed 7%. On the basis of the data obtained dose characteristics of electrons, muons and photons for standard geometry are estimated. For all types of irradiation the maximum value of specific equivalent dose, nremxcm 2 /part. is presented. It is shown that published values of specific equivalent dose of electron radiation are higher in all the investigated energy range from 0.2 to 40 GeV, and for muon radiation a good agreement with the present experiment is observed. The highly precise results obtained which cover the wide dynamic range according to the energy of incident particles can serve as the basis for reconsidering the existing recommendations for dose characteristics of electron radiation [ru

  5. Out‐of‐field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators

    Science.gov (United States)

    Cardenas, Carlos E.; Nitsch, Paige L.; Kudchadker, Rajat J.; Howell, Rebecca M.

    2016-01-01

    Out‐of‐field doses from radiotherapy can cause harmful side effects or eventually lead to secondary cancers. Scattered doses outside the applicator field, neutron source strength values, and neutron dose equivalents have not been broadly investigated for high‐energy electron beams. To better understand the extent of these exposures, we measured out‐of‐field dose characteristics of electron applicators for high‐energy electron beams on two Varian 21iXs, a Varian TrueBeam, and an Elekta Versa HD operating at various energy levels. Out‐of‐field dose profiles and percent depth‐dose curves were measured in a Wellhofer water phantom using a Farmer ion chamber. Neutron dose was assessed using a combination of moderator buckets and gold activation foils placed on the treatment couch at various locations in the patient plane on both the Varian 21iX and Elekta Versa HD linear accelerators. Our findings showed that out‐of‐field electron doses were highest for the highest electron energies. These doses typically decreased with increasing distance from the field edge but showed substantial increases over some distance ranges. The Elekta linear accelerator had higher electron out‐of‐field doses than the Varian units examined, and the Elekta dose profiles exhibited a second dose peak about 20 to 30 cm from central‐axis, which was found to be higher than typical out‐of‐field doses from photon beams. Electron doses decreased sharply with depth before becoming nearly constant; the dose was found to decrease to a depth of approximately E(MeV)/4 in cm. With respect to neutron dosimetry, Q values and neutron dose equivalents increased with electron beam energy. Neutron contamination from electron beams was found to be much lower than that from photon beams. Even though the neutron dose equivalent for electron beams represented a small portion of neutron doses observed under photon beams, neutron doses from electron beams may need to be considered for

  6. Out-of-field doses and neutron dose equivalents for electron beams from modern Varian and Elekta linear accelerators.

    Science.gov (United States)

    Cardenas, Carlos E; Nitsch, Paige L; Kudchadker, Rajat J; Howell, Rebecca M; Kry, Stephen F

    2016-07-08

    Out-of-field doses from radiotherapy can cause harmful side effects or eventually lead to secondary cancers. Scattered doses outside the applicator field, neutron source strength values, and neutron dose equivalents have not been broadly investigated for high-energy electron beams. To better understand the extent of these exposures, we measured out-of-field dose characteristics of electron applicators for high-energy electron beams on two Varian 21iXs, a Varian TrueBeam, and an Elekta Versa HD operating at various energy levels. Out-of-field dose profiles and percent depth-dose curves were measured in a Wellhofer water phantom using a Farmer ion chamber. Neutron dose was assessed using a combination of moderator buckets and gold activation foils placed on the treatment couch at various locations in the patient plane on both the Varian 21iX and Elekta Versa HD linear accelerators. Our findings showed that out-of-field electron doses were highest for the highest electron energies. These doses typically decreased with increasing distance from the field edge but showed substantial increases over some distance ranges. The Elekta linear accelerator had higher electron out-of-field doses than the Varian units examined, and the Elekta dose profiles exhibited a second dose peak about 20 to 30 cm from central-axis, which was found to be higher than typical out-of-field doses from photon beams. Electron doses decreased sharply with depth before becoming nearly constant; the dose was found to decrease to a depth of approximately E(MeV)/4 in cm. With respect to neutron dosimetry, Q values and neutron dose equivalents increased with electron beam energy. Neutron contamination from electron beams was found to be much lower than that from photon beams. Even though the neutron dose equivalent for electron beams represented a small portion of neutron doses observed under photon beams, neutron doses from electron beams may need to be considered for special cases.

  7. Personal dose equivalent conversion coefficients for electrons to 1 Ge V.

    Science.gov (United States)

    Veinot, K G; Hertel, N E

    2012-04-01

    In a previous paper, conversion coefficients for the personal dose equivalent, H(p)(d), for photons were reported. This note reports values for electrons calculated using similar techniques. The personal dose equivalent is the quantity used to approximate the protection quantity effective dose when performing personal dosemeter calibrations and in practice the personal dose equivalent is determined using a 30×30×15 cm slab-type phantom. Conversion coefficients to 1 GeV have been calculated for H(p)(10), H(p)(3) and H(p)(0.07) in the recommended slab phantom. Although the conversion coefficients were determined for discrete incident energies, analytical fits of the conversion coefficients over the energy range are provided using a similar formulation as in the photon results previously reported. The conversion coefficients for the personal dose equivalent are compared with the appropriate protection quantity, calculated according to the recommendations of the latest International Commission on Radiological Protection guidance. Effects of eyewear on H(p)(3) are also discussed.

  8. Calculations of the photon dose behind concrete shielding of high energy proton accelerators

    International Nuclear Information System (INIS)

    Dworak, D.; Tesch, K.; Zazula, J.M.

    1992-02-01

    The photon dose per primary beam proton behind lateral concrete shieldings was calculated by using an extension of the Monte Carlo particle shower code FLUKA. The following photon-producing processes were taken into account: capture of thermal neutrons, deexcitation of nuclei after nuclear evaporation, inelastic neutron scattering and nuclear reactions below 140 MeV, as well as photons from electromagnetic cascades. The obtained ratio of the photon dose to the neutron dose equivalent varies from 8% to 20% and it well compares with measurements performed recently at DESY giving a mean ratio of 14%. (orig.)

  9. Verification of an effective dose equivalent model for neutrons

    International Nuclear Information System (INIS)

    Tanner, J.E.; Piper, R.K.; Leonowich, J.A.; Faust, L.G.

    1992-01-01

    Since the effective dose equivalent, based on the weighted sum of organ dose equivalents, is not a directly measurable quantity, it must be estimated with the assistance of computer modelling techniques and a knowledge of the incident radiation field. Although extreme accuracy is not necessary for radiation protection purposes, a few well chosen measurements are required to confirm the theoretical models. Neutron doses and dose equivalents were measured in a RANDO phantom at specific locations using thermoluminescence dosemeters, etched track dosemeters, and a 1.27 cm (1/2 in) tissue-equivalent proportional counter. The phantom was exposed to a bare and a D 2 O-moderated 252 Cf neutron source at the Pacific Northwest Laboratory's Low Scatter Facility. The Monte Carlo code MCNP with the MIRD-V mathematical phantom was used to model the human body and to calculate the organ doses and dose equivalents. The experimental methods are described and the results of the measurements are compared with the calculations. (author)

  10. Verification of an effective dose equivalent model for neutrons

    International Nuclear Information System (INIS)

    Tanner, J.E.; Piper, R.K.; Leonowich, J.A.; Faust, L.G.

    1991-10-01

    Since the effective dose equivalent, based on the weighted sum of organ dose equivalents, is not a directly measurable quantity, it must be estimated with the assistance of computer modeling techniques and a knowledge of the radiation field. Although extreme accuracy is not necessary for radiation protection purposes, a few well-chosen measurements are required to confirm the theoretical models. Neutron measurements were performed in a RANDO phantom using thermoluminescent dosemeters, track etch dosemeters, and a 1/2-in. (1.27-cm) tissue equivalent proportional counter in order to estimate neutron doses and dose equivalents within the phantom at specific locations. The phantom was exposed to bare and D 2 O-moderated 252 Cf neutrons at the Pacific Northwest Laboratory's Low Scatter Facility. The Monte Carlo code MCNP with the MIRD-V mathematical phantom was used to model the human body and calculate organ doses and dose equivalents. The experimental methods are described and the results of the measurements are compared to the calculations. 8 refs., 3 figs., 3 tabs

  11. Dose determination algorithms for a nearly tissue equivalent multi-element thermoluminescent dosimeter

    International Nuclear Information System (INIS)

    Moscovitch, M.; Chamberlain, J.; Velbeck, K.J.

    1988-01-01

    In a continuing effort to develop dosimetric systems that will enable reliable interpretation of dosimeter readings in terms of the absorbed dose or dose-equivalent, a new multi-element TL dosimeter assembly for Beta and Gamma dose monitoring has been designed. The radiation-sensitive volumes are four LiF-TLD elements, each covered by its own unique filter. For media-matching, care has been taken to employ nearly tissue equivalent filters of thicknesses of 1000 mg/cm 2 and 300 mg/cm 2 for deep dose and dose to the lens-of-the-eye measurements respectively. Only one metal filter (Cu) is employed to provide low energy photon discrimination. A Thin TL element (0.09 mm thick) is located behind an open window designed to improve the energy under-response to low energy beta rays and to provide closer estimate of the shallow dose equivalent. The deep and shallow dose equivalents are derived from the correlation of the response of the various TL elements to the above quantities through computations based on previously defined relationships obtained from experimental results. The theoretical formalization for the dose calculation algorithms is described in detail, and provides a useful methodology which can be applied to different tissue-equivalent dosimeter assemblies. Experimental data has been obtained by performing irradiation according to the specifications established by DOELAP, using 27 types of pure and mixed radiation fields including Cs-137 gamma rays, low energy photons down to 20 keV, Sr/Y-90, Uranium, and Tl-204 beta particles

  12. Evaluation of effective dose equivalent from environmental gamma rays

    International Nuclear Information System (INIS)

    Saito, K.; Tsutsumi, M.; Moriuchi, S.; Petoussi, N.; Zankl, M.; Veit, R.; Jacob, P.; Drexler, G.

    1991-01-01

    Organ doses and effective dose equivalents for environmental gamma rays were calculated using human phantoms and Monte Carlo methods accounting rigorously the environmental gamma ray fields. It was suggested that body weight is the dominant factor to determine organ doses. The weight function expressing organ doses was introduced. Using this function, the variation in organ doses due to several physical factors were investigated. A detector having gamma-ray response similar to that of human bodies has been developed using a NaI(Tl) scintillator. (author)

  13. Measurement of absorbed dose with a bone-equivalent extrapolation chamber

    International Nuclear Information System (INIS)

    DeBlois, Francois; Abdel-Rahman, Wamied; Seuntjens, Jan P.; Podgorsak, Ervin B.

    2002-01-01

    A hybrid phantom-embedded extrapolation chamber (PEEC) made of Solid Water trade mark sign and bone-equivalent material was used for determining absorbed dose in a bone-equivalent phantom irradiated with clinical radiation beams (cobalt-60 gamma rays; 6 and 18 MV x rays; and 9 and 15 MeV electrons). The dose was determined with the Spencer-Attix cavity theory, using ionization gradient measurements and an indirect determination of the chamber air-mass through measurements of chamber capacitance. The collected charge was corrected for ionic recombination and diffusion in the chamber air volume following the standard two-voltage technique. Due to the hybrid chamber design, correction factors accounting for scatter deficit and electrode composition were determined and applied in the dose equation to obtain absorbed dose in bone for the equivalent homogeneous bone phantom. Correction factors for graphite electrodes were calculated with Monte Carlo techniques and the calculated results were verified through relative air cavity dose measurements for three different polarizing electrode materials: graphite, steel, and brass in conjunction with a graphite collecting electrode. Scatter deficit, due mainly to loss of lateral scatter in the hybrid chamber, reduces the dose to the air cavity in the hybrid PEEC in comparison with full bone PEEC by 0.7% to ∼2% depending on beam quality and energy. In megavoltage photon and electron beams, graphite electrodes do not affect the dose measurement in the Solid Water trade mark sign PEEC but decrease the cavity dose by up to 5% in the bone-equivalent PEEC even for very thin graphite electrodes (<0.0025 cm). In conjunction with appropriate correction factors determined with Monte Carlo techniques, the uncalibrated hybrid PEEC can be used for measuring absorbed dose in bone material to within 2% for high-energy photon and electron beams

  14. Evaluation of fluence to dose equivalent conversion factors for high energy radiations, (1)

    International Nuclear Information System (INIS)

    Sato, Osamu; Uehara, Takashi; Yoshizawa, Nobuaki; Iwai, Satoshi; Tanaka, Shun-ichi.

    1992-09-01

    Computer code system and basic data have been investigated for evaluating fluence to dose equivalent conversion factors for photons and neutrons up to 10 GeV. The present work suggested that the conversion factors would be obtained by incorporating effective quality factors of charged particles into the HERMES (High Energy Radiation Monte Carlo Elaborate System) code system. The effective quality factors for charged particles were calculated on the basis of the Q-L relationships specified in the ICRP Publication-60. (author)

  15. Testing of the analytical anisotropic algorithm for photon dose calculation

    International Nuclear Information System (INIS)

    Esch, Ann van; Tillikainen, Laura; Pyykkonen, Jukka; Tenhunen, Mikko; Helminen, Hannu; Siljamaeki, Sami; Alakuijala, Jyrki; Paiusco, Marta; Iori, Mauro; Huyskens, Dominique P.

    2006-01-01

    The analytical anisotropic algorithm (AAA) was implemented in the Eclipse (Varian Medical Systems) treatment planning system to replace the single pencil beam (SPB) algorithm for the calculation of dose distributions for photon beams. AAA was developed to improve the dose calculation accuracy, especially in heterogeneous media. The total dose deposition is calculated as the superposition of the dose deposited by two photon sources (primary and secondary) and by an electron contamination source. The photon dose is calculated as a three-dimensional convolution of Monte-Carlo precalculated scatter kernels, scaled according to the electron density matrix. For the configuration of AAA, an optimization algorithm determines the parameters characterizing the multiple source model by optimizing the agreement between the calculated and measured depth dose curves and profiles for the basic beam data. We have combined the acceptance tests obtained in three different departments for 6, 15, and 18 MV photon beams. The accuracy of AAA was tested for different field sizes (symmetric and asymmetric) for open fields, wedged fields, and static and dynamic multileaf collimation fields. Depth dose behavior at different source-to-phantom distances was investigated. Measurements were performed on homogeneous, water equivalent phantoms, on simple phantoms containing cork inhomogeneities, and on the thorax of an anthropomorphic phantom. Comparisons were made among measurements, AAA, and SPB calculations. The optimization procedure for the configuration of the algorithm was successful in reproducing the basic beam data with an overall accuracy of 3%, 1 mm in the build-up region, and 1%, 1 mm elsewhere. Testing of the algorithm in more clinical setups showed comparable results for depth dose curves, profiles, and monitor units of symmetric open and wedged beams below d max . The electron contamination model was found to be suboptimal to model the dose around d max , especially for physical

  16. The experimental method for neutron dose-equivalent detection

    International Nuclear Information System (INIS)

    Ji Changsong

    1992-01-01

    A new method, for getting neutron dose-equivalent Cd rode absorption method is described. The method adopts Cd-rode-swarm buck absorption, which greatly improved the neutron sensitivity and simplified the adjustment method. By this method, the author has developed BH3105 model neutron dose equivalent meter, the sensitivity of this instrument reach 10 cps/μSvh -1 . γ-ray depression rate reaches 4000:1, the measurement range is 0.1 μSv/h-10 6 μSv/h. The energy response is good (from thermal neutron-14 MeV neutron), this instrument can be used to measure the dose equivalent of the neutron areas

  17. Is the dose equivalent index a quantity to be measured

    International Nuclear Information System (INIS)

    Wagner, S.R.

    1980-01-01

    ICRP introduced the concept of Effective Dose Equivalent H(sub)E and fixed the basic limits of radiation exposure in terms of H(sub)I. As H(sub)E cannot be measured, ICRP stated that with external exposure to penetrating radiation the limitation of the Dose Equivalent Index H(sub)I would afford at least as good a level of protection. However, difficulties arise in measuring H(sub)I and in calibrating instruments in terms of H(sub)I, since the height and location of the dose equivalent maximum in the sphere which is the phantom used in the definition of H(sub)I, depend on the energy and the angular distribution of the incident radiation. That is, H(sub)I is not an additive quantity relative to the partial H(sub)I(sub)i-values of the different energy and angular components. Hence, 1) the distribution of dose equivalent in the sphere must be measured in full for a determination of H(sub)I, and 2) it is not possible to calibrate an instrument which does not exhibit the scattering and absorption properties of the sphere, consistently for arbitrary radiation fields in terms of H(sub)I. Thus the calibration in an unidirectional beam would infer an uncertainty which may amount to a factor of up to 4. This would hardly be tolerable as a base for radiation protection provisions. An alternative is to introduce operational quantities which are additive, e.g. 1) the sum of maxima of the dose equivalent distributions in the sphere produced by different radiation components, and 2) the mean dose equivalent in the sphere. Their relation to H(sub)E for different types of radiation and consequences on secondary limits are discussed. (H.K.)

  18. Comparison of proton and photon dose distributions

    International Nuclear Information System (INIS)

    Goitein, Michael

    1995-01-01

    Recently, there has been considerable work, as yet largely theoretical, in developing ways to improve the dose distributions which can be achieved with x-rays. Foremost among these developments are the use of non-coplanar beam directions, the use of intensity-modulated beams, and the implementation of computer-controlled delivery of complex plans using new beam modifiers such as multi-leaf collimators and beam scanners. One way of improving the dose distributions which have been achieved with conventional radiations is to use protons, with their quite different physical characteristics but very similar radiobiological properties as compared with supervoltage x-rays. Some substantial experience has been gained in the use of protons which has confirmed clinically that better results have been obtained as a result of their better dose distributions. Indeed, it is fair to say that the advantages which protons have demonstrated are, in large part, responsible for the renewed interest in improving the dose distributions from all radiation modalities. So much better are the dose distributions which the new techniques, mentioned above, offer that there is the impression that, with their use, photons can deliver dose distributions as good as can be obtained with protons. In this paper, the extent of the possible improvement will be discussed. It will be suggested that the integral dose is relatively little affected by the treatment technique - so that the lower normal tissue doses which the new approaches offer is almost always at the price of delivering dose to a larger volume. Protons can be matched pencil beam for pencil beam with photons - and then almost always deliver substantially less dose outside the target volume. Ultimately, the clinical importance of the differences will have to decided by clinical trial

  19. Is the dose equivalent index a quantity to be measured

    International Nuclear Information System (INIS)

    Wagner, S.R.

    1980-01-01

    The following modifying factors are briefly considered in relation to the ambiguities and limitations of the Dose Equivalent Index: 1) Variations with time or of the movement of the exposed person 2) Irradiation geometry 3) Effect of radiation energy 4) Instrument performance and calibration, and other operational quantities. (U.K.)

  20. Technical background for shallow (skin) dose equivalent evaluations

    International Nuclear Information System (INIS)

    Ashley, J.C.; Turner, J.E.; Crawford, O.H.; Hamm, R.N.; Reaves, K.L.; McMahan, K.L.

    1991-01-01

    Department of Energy Order 5480.11 describes procedures for radiation protection for occupational workers. The revisions dealing with non-uniform exposure to the skin are the subject of this report. We describe measurements and analysis required to assess shallow (skin) dose equivalent from skin contamination. 6 refs., 4 tabs

  1. Personnel dose equivalent monitoring at SLAC using lithium-fluoride TLD's [thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Jenkins, T.M.; Busick, D.D.

    1987-03-01

    TLD's replaced film badges in the early 1970's for all dose equivalent monitoring, both neutron and photon, and for all locations at SLAC. The photon TLD's, composed of Li-7 loaded teflon discs, are calibrated using conventional gamma-ray sources; i.e., Co-60, Cs-137, etc. For these TLD's a nominal value of 1 nC/mrem is used, and is independent of source energy for 100 keV to 3 MeV. Since measured dose equivalents at SLAC are only a small fraction of the allowable levels, it was not deemed necessary to develop neutron dosimeters which would measure dose equivalent accurately for all possible neutron spectra. Today, wallet TLD's, composed of pairs of Li-7 and Li-6 discs, are used, with the Li-6 measuring only thermal neutrons; i.e., they aren't moderated in any way to make them sensitive to neutrons with energies greater than thermal. The assumption is made that there is a correlation between thermal neutron fluences and fast neutron fluences around the research area where almost all neutron doses (exclusive of sealed sources) are received. The calibration factor for these Li-6 TLD's is 1 nC/mrem of fast neutrons. The method of determining the validity of this calibration is the subject of this note. 4 refs., 9 figs., 1 tab

  2. Peripheral photon and neutron doses from prostate cancer external beam irradiation.

    Science.gov (United States)

    Bezak, Eva; Takam, Rundgham; Marcu, Loredana G

    2015-12-01

    Peripheral photon and neutron doses from external beam radiotherapy (EBRT) are associated with increased risk of carcinogenesis in the out-of-field organs; thus, dose estimations of secondary radiation are imperative. Peripheral photon and neutron doses from EBRT of prostate carcinoma were measured in Rando phantom. (6)LiF:Mg,Cu,P and (7)LiF:Mg,Cu,P glass-rod thermoluminescence dosemeters (TLDs) were inserted in slices of a Rando phantom followed by exposure to 80 Gy with 18-MV photon four-field 3D-CRT technique. The TLDs were calibrated using 6- and 18-MV X-ray beam. Neutron dose equivalents measured with CR-39 etch-track detectors were used to derive readout-to-neutron dose conversion factor for (6)LiF:Mg,Cu,P TLDs. Average neutron dose equivalents per 1 Gy of isocentre dose were 3.8±0.9 mSv Gy(-1) for thyroid and 7.0±5.4 mSv Gy(-1) for colon. For photons, the average dose equivalents per 1 Gy of isocentre dose were 0.2±0.1 mSv Gy(-1) for thyroid and 8.1±9.7 mSv Gy(-1) for colon. Paired (6)LiF:Mg,Cu,P and (7)LiF:Mg,Cu,P TLDs can be used to measure photon and neutron doses simultaneously. Organs in close proximity to target received larger doses from photons than those from neutrons whereas distally located organs received higher neutron versus photon dose. © The Author 2015. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  3. Dose conversion coefficients for high-energy photons, electrons, neutrons and protons

    CERN Document Server

    Sakamoto, Y; Sato, O; Tanaka, S I; Tsuda, S; Yamaguchi, Y; Yoshizawa, N

    2003-01-01

    In the International Commission on Radiological Protection (ICRP) 1990 Recommendations, radiation weighting factors were introduced in the place of quality factors, the tissue weighting factors were revised, and effective doses and equivalent doses of each tissues and organs were defined as the protection quantities. Dose conversion coefficients for photons, electrons and neutrons based on new ICRP recommendations were cited in the ICRP Publication 74, but the energy ranges of theses data were limited and there are no data for high energy radiations produced in accelerator facilities. For the purpose of designing the high intensity proton accelerator facilities at JAERI, the dose evaluation code system of high energy radiations based on the HERMES code was developed and the dose conversion coefficients of effective dose were evaluated for photons, neutrons and protons up to 10 GeV, and electrons up to 100 GeV. The dose conversion coefficients of effective dose equivalent were also evaluated using quality fact...

  4. A design of ambient dose equivalent dosimeter and its dosimetric performance

    International Nuclear Information System (INIS)

    Zhao Shian; Ou Xiangming; Li Kaibao

    1997-01-01

    Objective: To design an ambient dose equivalent dosimeter with digital display for radiation protection, which is based on the definition of the new operational radiation quantity for environmental monitoring-ambient dose equivalent recommended by the International Commission on Radiation Units and Measurements (ICRU) Report 39. Methods: Considering the energy response of the instrument, the inner wall of ionizing chamber is coated with gum graphite added with a bit of metal powder. Results: Using this chamber, measurement of H * (10) for photon radiation with unknown spectrum distribution is possible in the energy range from 47 keV to 230 keV with an uncertainty of better than 5%. The configuration, technology and dosimetric performance of the chamber and automatic functions of the reader are presented. Conclusion: The ambient dose equivalent dosimeter can be used as not only a working reference dosimeter, but also a field dosimeter for radiation protection because the readings are expressed directly in ambient dose equivalent and averaged automatically in the period of measurement. Also, its power is supplied by battery for the portable purpose and the readings are displayed on the screen with light-background for dim field

  5. Antipsychotic dose equivalents and dose-years: a standardized method for comparing exposure to different drugs.

    Science.gov (United States)

    Andreasen, Nancy C; Pressler, Marcus; Nopoulos, Peg; Miller, Del; Ho, Beng-Choon

    2010-02-01

    A standardized quantitative method for comparing dosages of different drugs is a useful tool for designing clinical trials and for examining the effects of long-term medication side effects such as tardive dyskinesia. Such a method requires establishing dose equivalents. An expert consensus group has published charts of equivalent doses for various antipsychotic medications for first- and second-generation medications. These charts were used in this study. Regression was used to compare each drug in the experts' charts to chlorpromazine and haloperidol and to create formulas for each relationship. The formulas were solved for chlorpromazine 100 mg and haloperidol 2 mg to derive new chlorpromazine and haloperidol equivalents. The formulas were incorporated into our definition of dose-years such that 100 mg/day of chlorpromazine equivalent or 2 mg/day of haloperidol equivalent taken for 1 year is equal to one dose-year. All comparisons to chlorpromazine and haloperidol were highly linear with R(2) values greater than .9. A power transformation further improved linearity. By deriving a unique formula that converts doses to chlorpromazine or haloperidol equivalents, we can compare otherwise dissimilar drugs. These equivalents can be multiplied by the time an individual has been on a given dose to derive a cumulative value measured in dose-years in the form of (chlorpromazine equivalent in mg) x (time on dose measured in years). After each dose has been converted to dose-years, the results can be summed to provide a cumulative quantitative measure of lifetime exposure. Copyright 2010 Society of Biological Psychiatry. Published by Elsevier Inc. All rights reserved.

  6. Determination of dose equivalent and risk in thorium cycle

    International Nuclear Information System (INIS)

    Ney, C.L.V.N.

    1988-01-01

    In these report are presented the calculations of dose equivalent and risk, utilizing the dosimetric model described in publication 30 of the International Comission on Radiological Protection. This information was obtained by the workers of the thorium cycle, employed at the Praia and Santo Amaro Facilities, by assessing the quantity and concentration of thorium in the air. The samples and the number of measurements were established through design of experiments techniques, and the results were evaluated with the aid of variance analysis. The estimater of dose equivalent for internal and external radiation exposure and risk associated were compared with the maximum recommended limits. The results indicate the existence of operation areas whose values were above those limits, requiring so an improvement in the procedures and services in order to meet the requirements of the radiological protetion. (author) [pt

  7. Development of a neutron personal dose equivalent detector

    International Nuclear Information System (INIS)

    Tsujimura, N.; Yoshida, T.; Takada, C.; Momose, T.; Nunomiya, T.; Aoyama, K.

    2007-01-01

    A new neutron-measuring instrument that is intended to measure a neutron personal dose equivalent, H p (10) was developed. This instrument is composed of two parts: (1) a conventional moderator-based neutron dose equivalent meter and (2) a neutron shield made of borated polyethylene, which covers a backward hemisphere to adjust the angular dependence. The whole design was determined on the basis of MCNP calculations so as to have response characteristics that would generally match both the energy and angular dependencies of H p (10). This new instrument will be a great help in assessing the reference values of neutron H p (10) during field testing of personal neutron dosemeters in workplaces and also in interpreting their readings. (authors)

  8. Dose distribution around ion track in tissue equivalent material

    International Nuclear Information System (INIS)

    Zhang Wenzhong; Guo Yong; Luo Yisheng

    2007-01-01

    Objective: To study the energy deposition micro-specialty of ions in body-tissue or tissue equivalent material (TEM). Methods: The water vapor was determined as the tissue equivalent material, based on the analysis to the body-tissue, and Monte Carlo method was used to simulate the behavior of proton in the tissue equivalent material. Some features of the energy deposition micro-specialty of ion in tissue equivalent material were obtained through the analysis to the data from calculation. Results: The ion will give the energy by the way of excitation and ionization in material, then the secondary electrons will be generated in the progress of ionization, these electron will finished ions energy deposition progress. When ions deposited their energy, large amount energy will be in the core of tracks, and secondary electrons will devote its' energy around ion track, the ion dose distribution is then formed in TEM. Conclusions: To know biological effects of radiation , the research to dose distribution of ions is of importance(significance). (authors)

  9. Dose conversion coefficients for high-energy photons, electrons, neutrons and protons

    International Nuclear Information System (INIS)

    Sakamoto, Yukio

    2005-01-01

    Dose conversion coefficients for photons, electrons and neutrons based on new ICRP recommendations were cited in the ICRP Publication 74, but the energy ranges of these data were limited and there are no data for high energy radiations produced in accelerator facilities. For the purpose of designing the high intensity proton accelerator facilities at JAERI, the dose evaluation code system of high energy radiations based on the HERMES code was developed and the dose conversion coefficients of effective dose were evaluated for photons, neutrons and protons up to 10 GeV, and electrons up to 100 GeV. The dose conversion coefficients of effective dose equivalent were also evaluated using quality factors to consider the consistency between radiation weighting factors and Q-L relationship. The effective dose conversion coefficients obtained in this work were in good agreement with those recently evaluated by using FLUKA code for photons and electrons with all energies, and neutrons and protons below 500 MeV. There were some discrepancy between two data owing to the difference of cross sections in the nuclear reaction models. The dose conversion coefficients of effective dose equivalents for high energy radiations based on Q-L relation in ICRP Publication 60 were evaluated only in this work. The previous comparison between effective dose and effective dose equivalent made it clear that the radiation weighting factors for high energy neutrons and protons were overestimated and the modification was required. (author)

  10. Determination of the dose equivalents due to neutrons produced during therapeutic irradiations with a Varian CLINAC 2500

    International Nuclear Information System (INIS)

    Carrillo, Ricardo E.

    1991-01-01

    This experiment it was designed to quantify that so important it is the dose equivalent deposited by the neutron flow that is generated by photonuclear reactions during therapeutic irradiations with X rays of produced high-energy for an accelerator Varian CLINAC 2500. This accelerator type is routinely used in the Department of Radiotherapy of the Hospital of the University of Wisconsin, E.U. The equivalent dose was measured in diverse towns of the room of irradiations using the activation of thin sheets of gold put in the center of plastic recipients full with water. In general, the recipients were 1 m or more than the floor and at distances still bigger than the walls. The irradiations were made using photons with the highest energy that you can select with this team - 24 MeV. The due equivalent dose to neutrons taken place here by the energy photons used they were measured and reported. (author)

  11. Measurement of neutron dose equivalent outside and inside of the treatment vault of GRID therapy

    Energy Technology Data Exchange (ETDEWEB)

    Wang, Xudong; Charlton, Michael A.; Esquivel, Carlos; Eng, Tony Y.; Li, Ying; Papanikolaou, Nikos [University of Texas Health Science Center, San Antonio, Texas 78229 (United States)

    2013-09-15

    Purpose: To evaluate the neutron and photon dose equivalent rates at the treatment vault entrance (H{sub n,D} and H{sub G}), and to study the secondary radiation to the patient in GRID therapy. The radiation activation on the grid was studied.Methods: A Varian Clinac 23EX accelerator was working at 18 MV mode with a grid manufactured by .decimal, Inc. The H{sub n,D} and H{sub G} were measured using an Andersson–Braun neutron REM meter, and a Geiger Müller counter. The radiation activation on the grid was measured after the irradiation with an ion chamber γ-ray survey meter. The secondary radiation dose equivalent to patient was evaluated by etched track detectors and OSL detectors on a RANDO{sup ®} phantom.Results: Within the measurement uncertainty, there is no significant difference between the H{sub n,D} and H{sub G} with and without a grid. However, the neutron dose equivalent to the patient with the grid is, on average, 35.3% lower than that without the grid when using the same field size and the same amount of monitor unit. The photon dose equivalent to the patient with the grid is, on average, 44.9% lower. The measured average half-life of the radiation activation in the grid is 12.0 (±0.9) min. The activation can be categorized into a fast decay component and a slow decay component with half-lives of 3.4 (±1.6) min and 15.3 (±4.0) min, respectively. There was no detectable radioactive contamination found on the surface of the grid through a wipe test.Conclusions: This work indicates that there is no significant change of the H{sub n,D} and H{sub G} in GRID therapy, compared with a conventional external beam therapy. However, the neutron and scattered photon dose equivalent to the patient decrease dramatically with the grid and can be clinical irrelevant. Meanwhile, the users of a grid should be aware of the possible high dose to the radiation worker from the radiation activation on the surface of the grid. A delay in handling the grid after the beam

  12. Evaluation of equivalent doses in 18F PET/CT using the Monte Carlo method with MCNPX code

    International Nuclear Information System (INIS)

    Belinato, Walmir; Santos, William Souza; Perini, Ana Paula; Neves, Lucio Pereira; Souza, Divanizia N.

    2017-01-01

    The present work used the Monte Carlo method (MMC), specifically the Monte Carlo NParticle - MCNPX, to simulate the interaction of radiation involving photons and particles, such as positrons and electrons, with virtual adult anthropomorphic simulators on PET / CT scans and to determine absorbed and equivalent doses in adult male and female patients

  13. Improving the equivalent-photon approximation in electron-positron collisions

    CERN Document Server

    Schuler, G A

    1996-01-01

    The validity of the equivalent-photon approximation for two-photon processes in electron--positron collisions is critically examined. Commonly used forms to describe hadronic two-photon production are shown to lead to sizeable errors. An improved two-photon luminosity function is presented, which includes beyond-leading-logarithmic effects and scalar-photon contributions. Comparisons of various approximate expressions with the exact calculation in the case of the total hadronic cross section are given. Furthermore, effects of the poorly known low-Q2 behaviour of the virtual hadronic cross sections are discussed.

  14. BH3105 type neutron dose equivalent meter of high sensitivity

    International Nuclear Information System (INIS)

    Ji Changsong; Zhang Enshan; Yang Jianfeng; Zhang Hong; Huang Jiling

    1995-10-01

    It is noted that to design a neutron dose meter of high sensitivity is almost impossible in the frame of traditional designing principle--'absorption net principle'. Based on a newly proposed principle of obtaining neutron dose equi-biological effect adjustment--' absorption stick principle', a brand-new neutron dose-equivalent meter with high neutron sensitivity BH3105 has been developed. Its sensitivity reaches 10 cps/(μSv·h -1 ), which is 18∼40 times higher than one of foreign products of the same kind and is 10 4 times higher than that of domestic FJ342 neutron rem-meter. BH3105 has a measurement range from 0.1μSv/h to 1 Sv/h which is 1 or 2 orders wider than that of the other's. It has the advanced properties of gamma-resistance, energy response, orientation, etc. (6 tabs., 5 figs.)

  15. Angular dependence of depth doses in a tissue slab irradiated with monoenergetic photons

    International Nuclear Information System (INIS)

    Till, E.; Zankl, M.; Drexler, G.

    1995-12-01

    This report presents dose equivalents from external photon irradiation, normalised to air kerma free in air, on the central axis of a cuboid slab of ICRU tissue for various depths, photon energies and angles of beam incidence. The data were calculated by a Monte Carlo method using an idealised planar parallel source of monoenergetic photons. The data presented here aim at facilitating the calibration of individual dosimeters; they provide also an estimate of the quantity 'personal dose equivalent' defined by the ICRU. A detailed evaluation of the dependence of the calculated conversion coefficients on depth in the slab, photon energy and angle of incidence is given. A comparison with published measured an calculated values of angular dependence factors is made. (orig.)

  16. Collective effective dose equivalent, population doses and risk estimates from occupational exposures in Japan

    International Nuclear Information System (INIS)

    Maruyama, Takashi; Nishizawa, Kanae; Kumamoto, Yoshikazu; Iwai, Kazuo; Mase, Naomichi.

    1993-01-01

    Collective dose equivalent and population dose from occupational exposures in Japan, 1988 were estimated on the basis of a nationwide survey. The survey was conducted on annual collective dose equivalents by sex, age group and type of radiation work for about 0.21 million workers except for the workers in nuclear power stations. The data on the workers in nuclear power stations were obtained from the official report of the Japan Nuclear Safety Commission. The total number of workers including nuclear power stations was estimated to be about 0.26 million. Radiation works were subdivided as follows: medical works including dental; non-atomic energy industry; research and education; atomic energy industry and nuclear power station. For the determination of effective dose equivalent and population dose, organ or tissue doses were measured with a phantom experiment. The resultant doses were compared with the doses previously calculated using a chord length technique and with data from ICRP publications. The annual collective effective dose equivalent were estimated to be about 21.94 person·Sv for medical workers, 7.73 person·Sv for industrial workers, 0.75 person·Sv for research and educational workers, 2.48 person·Sv for atomic energy industry and 84.4 person ·Sv for workers in nuclear power station. The population doses were calculated to be about 1.07 Sv for genetically significant dose, 0.89 Sv for leukemia significant dose and 0.42 Sv for malignant significant dose. The population risks were estimated using these population doses. (author)

  17. Bone marrow equivalent prompt dose from two common fallout scenarios

    International Nuclear Information System (INIS)

    Morris, M.D.; Jones, T.D.; Young, R.W.

    1994-01-01

    A cell-kinetics model for radiation-induced myelopoiesis has been derived for mice, rats, dogs, sheep, swine, and burros. The model was extended to humans after extensive comparisons with molecular and cellular data from biological experiments and an assortment of predictive/validation tests on animal mortality, cell survival, and cellular repopulation following irradiations. One advantage of the model is that any complex pattern of protracted irradiation can be equated to its equivalent prompt dose. Severity of biological response depends upon target-organ dose, dose rate, and dose fractionation. Epidemiological and animal data are best suited for exposures given in brief periods of time. To use those data to assess risk from protracted human exposures, it is obligatory to model molecular repair and compensatory proliferation in terms of prompt dose. Although the model is somewhat complex both mathematically and biologically, this note describes simple numerical approximations for two common exposure scenarios. Both approximations are easily evaluated on a simple pocket calculator by a health physicist or emergency management officer. 12 refs., 5 figs

  18. Equivalent dose determination in foraminifera: analytical description of the CO2--signal dose-response curve

    International Nuclear Information System (INIS)

    Hoffmann, D.; Woda, C.; Mangini, A.

    2003-01-01

    The dose-response of the CO 2 - signal (g=2.0006) in foraminifera with ages between 19 and 300 ka is investigated. The sum of two exponential saturation functions is an adequate function to describe the dose-response curve up to an additional dose of 8000 Gy. It yields excellent dating results but requires an artificial doses of at least 5000 Gy. For small additional doses of about 500 Gy the single exponential saturation function can be used to calculate a reliable equivalent dose D E , although it does not describ the dose-response for higher doses. The CO 2 - -signal dose-response indicates that the signal has two components of which one is less stable than the other

  19. Experimental method research on neutron equal dose-equivalent detection

    International Nuclear Information System (INIS)

    Ji Changsong

    1995-10-01

    The design principles of neutron dose-equivalent meter for neutron biological equi-effect detection are studied. Two traditional principles 'absorption net principle' and 'multi-detector principle' are discussed, and on the basis of which a new theoretical principle for neutron biological equi-effect detection--'absorption stick principle' has been put forward to place high hope on both increasing neutron sensitivity of this type of meters and overcoming the shortages of the two traditional methods. In accordance with this new principle a brand-new model of neutron dose-equivalent meter BH3105 has been developed. Its neutron sensitivity reaches 10 cps/(μSv·h -1 ), 18∼40 times higher than that of all the same kinds of meters 0.23∼0.56 cps/(μSv·h -1 ), available today at home and abroad and the specifications of the newly developed meter reach or surpass the levels of the same kind of meters. Therefore the new theoretical principle of neutron biological equi-effect detection--'absorption stick principle' is proved to be scientific, advanced and useful by experiments. (3 refs., 3 figs., 2 tabs.)

  20. Response of radiation monitors for ambient dose equivalent, H*(10)

    International Nuclear Information System (INIS)

    Grecco, Claudio Henrique dos Santos

    2001-01-01

    Radiation monitors are used all over the world to evaluate if places with presence of ionising radiation present safe conditions for people. Radiation monitors should be tested according to international or national standards in order to be qualified for use. This work describes a methodology and procedures to evaluate the energy and angular responses of any radiation monitor for ambient dose equivalent, H*(10), according to the recommendations of ISO and IEC standards. The methodology and the procedures were applied to the Monitor Inteligente de Radiacao MIR 7026, developed by the Instituto em Engenharia Nuclear (IEN), to evaluate and to adjust its response for H*(10), characterizing it as an ambient dose equivalent meter. The tests were performed at the Laboratorio Nacional de Metrologia das Radiacoes Ionizantes (LNMRI), at Instituto de Radioprotecao e Dosimetria (IRD), and results showed that the Monitor Inteligente de Radiacao MIR 7026 can be used as an EI*(10) meter, in accordance to the IEC 60846 standard requirements. The overall estimated uncertainty for the determination of the MIR 7026 response, in all radiation qualities used in this work, was 4,5 % to a 95 % confidence limit. (author)

  1. Calculation of neutron fluence to dose equivalent conversion coefficients using GEANT4

    International Nuclear Information System (INIS)

    Ribeiro, Rosane M.; Santos, Denison de S.; Queiroz Filho, Pedro P. de; Mauricio, CLaudia L.P.; Silva, Livia K. da; Pessanha, Paula R.

    2014-01-01

    Fluence to dose equivalent conversion coefficients provide the basis for the calculation of area and personal monitors. Recently, the ICRP has started a revision of these coefficients, including new Monte Carlo codes for benchmarking. So far, little information is available about neutron transport below 10 MeV in tissue-equivalent (TE) material performed with Monte Carlo GEANT4 code. The objective of this work is to calculate neutron fluence to personal dose equivalent conversion coefficients, H p (10)/Φ, with GEANT4 code. The incidence of monoenergetic neutrons was simulated as an expanded and aligned field, with energies ranging between thermal neutrons to 10 MeV on the ICRU slab of dimension 30 x 30 x 15 cm 3 , composed of 76.2% of oxygen, 10.1% of hydrogen, 11.1% of carbon and 2.6% of nitrogen. For all incident energy, a cylindrical sensitive volume is placed at a depth of 10 mm, in the largest surface of the slab (30 x 30 cm 2 ). Physic process are included for neutrons, photons and charged particles, and calculations are made for neutrons and secondary particles which reach the sensitive volume. Results obtained are thus compared with values published in ICRP 74. Neutron fluence in the sensitive volume was calculated for benchmarking. The Monte Carlo GEANT4 code was found to be appropriate to calculate neutron doses at energies below 10 MeV correctly. (author)

  2. Committed equivalent organ doses and committed effective doses from intakes of radionuclides

    CERN Document Server

    Phipps, A W; Kendall, G M; Silk, T J; Stather, J W

    1991-01-01

    This report contains details of committed equivalent doses to individual organs for intakes by ingestion and inhalation of 1 mu m AMAD particles of 359 nuclides by infants aged 3 months, by children aged 1, 5, 10 and 15 years, and by adults. It complements NRPB-R245 which describes the changes which have taken place since the last NRPB compendium of dose per unit intake factors (dose coefficients) and gives summary tables. Information on the way committed doses increase with the integration period is given in NRPB-M289. The information given in these memoranda is also available as a microcomputer package - NRPB-SR245.

  3. Photon spectrometry for the determination of the dose-rate constant of low-energy photon-emitting brachytherapy sources

    International Nuclear Information System (INIS)

    Chen, Zhe Jay; Nath, Ravinder

    2007-01-01

    Accurate determination of dose-rate constant (Λ) for interstitial brachytherapy sources emitting low-energy photons (<50 keV) has remained a challenge in radiation dosimetry because of the lack of a suitable absolute dosimeter for accurate measurement of the dose rates near these sources. Indeed, a consensus value of Λ taken as the arithmetic mean of the dose-rate constants determined by different research groups and dosimetry techniques has to be used at present for each source model in order to minimize the uncertainties associated with individual determinations of Λ. Because the dosimetric properties of a source are fundamentally determined by the characteristics of the photons emitted by the source, a new technique based on photon spectrometry was developed in this work for the determination of dose-rate constant. The photon spectrometry technique utilized a high-resolution gamma-ray spectrometer to measure source-specific photon characteristics emitted by the low-energy sources and determine their dose-rate constants based on the measured photon-energy spectra and known dose-deposition properties of mono-energetic photons in water. This technique eliminates many of the difficulties arising from detector size, the energy dependence of detector sensitivity, and the use of non-water-equivalent solid phantoms in absolute dose rate measurements. It also circumvents the uncertainties that might be associated with the source modeling in Monte Carlo simulation techniques. It was shown that the estimated overall uncertainty of the photon spectrometry technique was less than 4%, which is significantly smaller than the reported 8-10% uncertainty associated with the current thermo-luminescent dosimetry technique. In addition, the photon spectrometry technique was found to be stable and quick in Λ determination after initial setup and calibration. A dose-rate constant can be determined in less than two hours for each source. These features make it ideal to determine

  4. Measurement of ambient dose equivalent H*(10) and directional dose equivalent H'(0.07) with pocket sized survey meters

    International Nuclear Information System (INIS)

    Iwatschenko, Michael

    2008-01-01

    Full text: In many parts of the world, predominantly in Europe, small sized survey meters based on Geiger-Mueller or proportional counters are widely used for dose rate and dose equivalent rate measurements, while in other regions, especially in the U.S., ionisation chambers are preferred for this task. This paper tries to shed some light on the likely reasons for these two diverging instrumental inclinations. Their respective strengths and weaknesses is analyzed in respect to energy response, dose rate measuring range, size, weight and susceptibility to environmental influences. Furthermore the response and limitations regarding the measurement of pulsed radiation (medical X-ray and CT-devices, accelerators, non-destructive testing) is discussed. A newly developed pocket size instrument based on a pan-cake Geiger-Mueller tube is used as an example to explain the capability and flexibility of modern survey meters. The RadEye B20 is a compact multi-purpose dose rate meter and contamination meter for alpha, beta, gamma and X-ray radiation. By virtue of carefully designed multi-layer gamma energy filters, H*(10) (deep dose) or H'(0,07) (shallow dose) measurements from 17 - 1300 keV can be performed. The instrument can even be worn in a belt holster, so that the impact to the mobility of the user is minimized. For emergency response purposes alpha and beta contamination can be discriminated using another optional filter; a simple sample changer adapter can extend the scope of application. Immediate and reproducible counter measurements, e.g. of smear tests can be performed locally. (author)

  5. The performance of low pressure tissue-equivalent chambers and a new method for parameterising the dose equivalent

    International Nuclear Information System (INIS)

    Eisen, Y.

    1986-01-01

    The performance of Rossi-type spherical tissue-equivalent chambers with equivalent diameters between 0.5 μm and 2 μm was tested experimentally using monoenergetic and polyenergetic neutron sources in the energy region of 10 keV to 14.5 MeV. In agreement with theoretical predictions both chambers failed to provide LET information at low neutron energies. A dose equivalent algorithm was derived that utilises the event distribution but does not attempt to correlate event size with LET. The algorithm was predicted theoretically and confirmed by experiment. The algorithm that was developed determines the neutron dose equivalent, from the data of the 0.5 μm chamber, to better than +-20% over the energy range of 30 keV to 14.5 MeV. The same algorithm also determines the dose equivalent from the data of the 2 μm chamber to better than +-20% over the energy range of 60 keV to 14.5 MeV. The efficiency of the chambers is 33 counts per μSv, or equivalently about 10 counts s -1 per mSv.h -1 . This efficiency enables the measurement of dose equivalent rates above 1 mSv.h -1 for an integration period of 3 s. Integrated dose equivalents can be measured as low as 1 μSv. (author)

  6. Evaluation of 1cm dose equivalent rate using a NaI(Tl) scintilation spectrometer

    International Nuclear Information System (INIS)

    Matsuda, Hideharu

    1990-01-01

    A method for evaluating 1 cm dose equivalent rates from a pulse height distribution obtained by a 76.2mmφ spherical NaI(Tl) scintillation spectrometer was described. Weak leakage radiation from nuclear facilities were also measured and dose equivalent conversion factor and effective energy of leakage radiation were evaluated from 1 cm dose equivalent rate and exposure rate. (author)

  7. 10 CFR 20.1208 - Dose equivalent to an embryo/fetus.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 1 2010-01-01 2010-01-01 false Dose equivalent to an embryo/fetus. 20.1208 Section 20... Limits § 20.1208 Dose equivalent to an embryo/fetus. (a) The licensee shall ensure that the dose equivalent to the embryo/fetus during the entire pregnancy, due to the occupational exposure of a declared...

  8. A single-aliquot OSL protocol using bracketing regenerative doses to accurately determine equivalent doses in quartz

    International Nuclear Information System (INIS)

    Folz, Elise; Mercier, Norbert

    1999-01-01

    In most cases, sediments show inherent heterogeneity in their luminescence behaviours and bleaching histories, and identical aliquots are not available: single-aliquot determination of the equivalent dose (ED) is then the approach of choice and the advantages of using regenerative protocols are outlined. Experiments on five laboratory bleached and dosed quartz samples, following the protocol described by Murray and Roberts (1998. Measurement of the equivalent dose in quartz using a regenerative-dose single aliquot protocol. Radiation Measurements 27, 171-184), showed the hazards of using a single regeneration dose: a 10% variation in the regenerative dose yielded some equivalent dose estimates that differed from the expected value by more than 5%. A protocol is proposed that allows the use of different regenerative doses to bracket the estimated equivalent dose. The measured ED is found to be in excellent agreement with the known value when the main regeneration dose is within 10% of the true equivalent dose

  9. A single-aliquot OSL protocol using bracketing regenerative doses to accurately determine equivalent doses in quartz

    CERN Document Server

    Folz, E

    1999-01-01

    In most cases, sediments show inherent heterogeneity in their luminescence behaviours and bleaching histories, and identical aliquots are not available: single-aliquot determination of the equivalent dose (ED) is then the approach of choice and the advantages of using regenerative protocols are outlined. Experiments on five laboratory bleached and dosed quartz samples, following the protocol described by Murray and Roberts (1998. Measurement of the equivalent dose in quartz using a regenerative-dose single aliquot protocol. Radiation Measurements 27, 171-184), showed the hazards of using a single regeneration dose: a 10% variation in the regenerative dose yielded some equivalent dose estimates that differed from the expected value by more than 5%. A protocol is proposed that allows the use of different regenerative doses to bracket the estimated equivalent dose. The measured ED is found to be in excellent agreement with the known value when the main regeneration dose is within 10% of the true equivalent dose.

  10. Total effective dose equivalent associated with fixed uranium surface contamination

    International Nuclear Information System (INIS)

    Bogard, J.S.; Hamm, R.N.; Ashley, J.C.; Turner, J.E.; England, C.A.; Swenson, D.E.; Brown, K.S.

    1997-04-01

    This report provides the technical basis for establishing a uranium fixed-contamination action level, a fixed uranium surface contamination level exceeding the total radioactivity values of Appendix D of Title 10, Code of Federal Regulations, part 835 (10CFR835), but below which the monitoring, posting, and control requirements for Radiological Areas are not required for the area of the contamination. An area of fixed uranium contamination between 1,000 dpm/100 cm 2 and that level corresponding to an annual total effective dose equivalent (TEDE) of 100 mrem requires only routine monitoring, posting to alert personnel of the contamination, and administrative control. The more extensive requirements for monitoring, posting, and control designated by 10CFR835 for Radiological Areas do not have to be applied for these intermediate fixed-contamination levels

  11. Determination of skin dose reduction by lead equivalent gloves

    International Nuclear Information System (INIS)

    Norriza Mohd Isa; Abd Aziz Mhd Ramli

    2006-01-01

    Radiation protective gloves are always used in medical facilities to protect radiation workers from unnecessary radiation exposure. A study on radiation protection gloves which are produced by local company had been performed by the Medical Physics Group, MINT. The gloves were made of lead equivalent material, as the attenuating element. The gloves were evaluated in term of the percentage of skin dose reduction by using a newly developed procedure and facilities in MINT. Attenuation measurements of the gloves had been carried out using direct beams and scattered radiations of different qualities. TLD rings were fitted on finger phantom; and water phantom were used in the measurement. The result were obtained and analysed based on data supplied by manufacturer. (Author)

  12. The effective dose equivalent from external and internal radiation

    International Nuclear Information System (INIS)

    Mattsson, Soeren

    1989-01-01

    The various sources of low-level ionizing radiation are discussed and compared in terms of mean effective dose equivalent to man. For the most nonoccupationally exposed individuals, natural sources given the dominating contribution to the effective dose equivalent. The size of this contribution is strongly dependent on human activities. Natural sources contribution on average 2.4 mSV per year, of which half is due to irradiation of lungs and airways from short lived radon daughters present in indoor air. In Sweden this radon daughter contribution is considerably higher and contributes a mean of 3 mSv per year, thus giving a total contribution from natural radiation of about 4 mSV per year. In extreme cases, radon daughter contributions of several hundreds of mSv per year may be reached. Medical exposure, mainly diagnostic X-rays, contributes 0.4-1 mSv per year both in Sweden and as a world average. The testing of nuclear weapons in the atmosphere has given 1-2 mSv to each person in the world as a mean. The contribution from the routine operation of nuclear reactors is insignificant. The reactor accident in Chernobyl resulted in widely varying exposures of the European population. The average for Sweden is estimated to be 0.1 mSv during the first year and about 1 mSv during a 50-year period. For groups of Swedes who eat a considerable amount of game this contribution will be 10 times higher, and for the Lapps who breed reindeer in the most contaminated areas, typical values of 20-70 mSv and extreme values of about 1 Sv may be reached in 50 years. This means that the Chernobyl reactor accident for several years will be their dominating source of irradiation

  13. Equivalent (uniform) square field sizes of flattening filter free photon beams

    Science.gov (United States)

    Lechner, Wolfgang; Kuess, Peter; Georg, Dietmar; Palmans, Hugo

    2017-10-01

    Various types of treatment units, such as CyberKnife, TomoTherapy and C-arm linear accelerators (LINACs) are operated using flattening filter free (FFF) photon beams. Their reference dosimetry, however, is currently based on codes of practice that provide data which were primarily developed and tested for high-energy photon beams with flattening filter (WFF). The aim of this work was to introduce equivalent uniform square field sizes of FFF beams to serve as a basis of a unified reference dosimetry procedure applicable to all aforementioned FFF machines. For this purpose, in-house determined experimental data together with published data of the ratio of doses at depths of 20 cm and 10 cm in water (D 20,10) were used to characterize the depth dose distribution of 6 and 10 MV WFF and FFF beams. These data were analyzed for field sizes ranging from 2  ×  2 cm2 to 40  ×  40 cm2. A scatter function that takes the lateral profiles of the individual beams into account was fitted to the experimental data. The lateral profiles of the WFF beams were assumed to be uniform, while those of the FFF beams were approximated using fourth or sixth order polynomials. The scatter functions of the FFF beams were recalculated using a uniform lateral profile (the same as the physical profile of the WFF beams), and are henceforth denoted as virtual uniform FFF beams (VUFFF). The field sizes of the VUFFF beams having the same scatter contribution as the corresponding FFF beams at a given field size were defined as the equivalent uniform square field (EQUSF) size. Data from four different LINACs with 18 different beams in total, as well as a CyberKnife beam, were analyzed. The average values of EQUSFs over all investigated LINACs of the conventional 10  ×  10 cm2 reference fields of 6 MV and 10 MV FFF beams for C-arm LINACs and machine-specific reference fields for CyberKnife and TomoTherapy were 9.5 cm, 9 cm, 5.0 cm and 6.5 cm respectively. The

  14. Dose inhomogeneities for photons and neutrons near interfaces

    International Nuclear Information System (INIS)

    Broerse, J. J.; Zoetelief, J.

    2004-01-01

    Perturbations of charged particle equilibrium (CPE) at interfaces of materials of different atomic composition can lead to considerable differences in the energy deposition by photons and neutrons. Specific examples of these interface perturbations are encountered during irradiation of body cavities and soft tissue adjacent to bone or metallic implants and irradiation of cells in monolayer on the bottom of culture dishes. Another example is the build-up of CPE at air-tissue interfaces, referred to in radiotherapy as the skin sparing effect. For photon irradiation excess production of secondary electrons in high-Z materials, such as glass, bone or gold, will induce appreciably higher doses and decreased cell survival compared to the equilibrium situation. The energy dissipation of fast neutrons in biological materials occurs through recoil protons, heavy recoil nuclei and products of nuclear reactions. Owing to the large contribution from recoil protons to the neutron kerma, the hydrogen content of the biological material mainly determines the energy deposition. For neutron irradiation of cells in monolayer, CPE can be established or deliberately avoided by mounting tissue-equivalent plastic or carbon discs in front of the cells, respectively. This approach makes it possible to distinguish the biological effects of the low- and high-LET radiation components. (authors)

  15. Photon and neutron doses of the personnel using moisture and density measurement devices

    Energy Technology Data Exchange (ETDEWEB)

    Carinou, E.; Papadomarkaki, E.; Tritakis, P.; Hourdakis, C.I.; Kamenopoulou, V. [Greek Atomic Energy Commission, Agia Paraskevi, Attiki, 60092 (Greece)

    2006-07-01

    The objective of this study is to present the evolution of the photon doses received by the workers who use mobile devices for measuring the moisture and the density in various materials and to estimate the neutron doses. The workers employed in more than 30 construction companies in Greece were 76 in 2004. The devices used for that purpose incorporate a {sup 137}Cs source for density measurements and an {sup 241}Am-Be source for moisture measurements of soil, asphalt or concrete. Photon and neutron measurements were performed occasionally during the on site inspections. The results of the measurements showed that the photon and neutron dose rates were not negligible. The workers were monitored for photon radiation using film badges (Kodak Type 2, Holder NRPB type) till the year 2000 and then TLD badges issued by the Greek Atomic Energy Commission (GAEC), on a monthly basis. Since the neutron dose rates measured by a rem-meter were not so high, no neutron dosemeters were issued for them. Their personal dose equivalent data for photons are kept in the National Dose Registry Information System (N.D.R.I.S.) in G.A.E.C. and were used for statistical analysis for the period from 1997 till 2004. As far as the neutrons are concerned, a Monte Carlo code was used to simulate the measuring devices and the working positions in order to calculate the neutron individual doses. (authors)

  16. Photon and neutron doses of the personnel using moisture and density measurement devices

    International Nuclear Information System (INIS)

    Carinou, E.; Papadomarkaki, E.; Tritakis, P.; Hourdakis, C.I.; Kamenopoulou, V.

    2006-01-01

    The objective of this study is to present the evolution of the photon doses received by the workers who use mobile devices for measuring the moisture and the density in various materials and to estimate the neutron doses. The workers employed in more than 30 construction companies in Greece were 76 in 2004. The devices used for that purpose incorporate a 137 Cs source for density measurements and an 241 Am-Be source for moisture measurements of soil, asphalt or concrete. Photon and neutron measurements were performed occasionally during the on site inspections. The results of the measurements showed that the photon and neutron dose rates were not negligible. The workers were monitored for photon radiation using film badges (Kodak Type 2, Holder NRPB type) till the year 2000 and then TLD badges issued by the Greek Atomic Energy Commission (GAEC), on a monthly basis. Since the neutron dose rates measured by a rem-meter were not so high, no neutron dosemeters were issued for them. Their personal dose equivalent data for photons are kept in the National Dose Registry Information System (N.D.R.I.S.) in G.A.E.C. and were used for statistical analysis for the period from 1997 till 2004. As far as the neutrons are concerned, a Monte Carlo code was used to simulate the measuring devices and the working positions in order to calculate the neutron individual doses. (authors)

  17. From physical dose constraints to equivalent uniform dose constraints in inverse radiotherapy planning

    International Nuclear Information System (INIS)

    Thieke, Christian; Bortfeld, Thomas; Niemierko, Andrzej; Nill, Simeon

    2003-01-01

    Optimization algorithms in inverse radiotherapy planning need information about the desired dose distribution. Usually the planner defines physical dose constraints for each structure of the treatment plan, either in form of minimum and maximum doses or as dose-volume constraints. The concept of equivalent uniform dose (EUD) was designed to describe dose distributions with a higher clinical relevance. In this paper, we present a method to consider the EUD as an optimization constraint by using the method of projections onto convex sets (POCS). In each iteration of the optimization loop, for the actual dose distribution of an organ that violates an EUD constraint a new dose distribution is calculated that satisfies the EUD constraint, leading to voxel-based physical dose constraints. The new dose distribution is found by projecting the current one onto the convex set of all dose distributions fulfilling the EUD constraint. The algorithm is easy to integrate into existing inverse planning systems, and it allows the planner to choose between physical and EUD constraints separately for each structure. A clinical case of a head and neck tumor is optimized using three different sets of constraints: physical constraints for all structures, physical constraints for the target and EUD constraints for the organs at risk, and EUD constraints for all structures. The results show that the POCS method converges stable and given EUD constraints are reached closely

  18. An analysis of depth dose characteristics of photon in water

    International Nuclear Information System (INIS)

    Buzdar, S.A.; Rao, M.A.; Nazir, A.

    2009-01-01

    Photon beam is most widely being used for radiation therapy. Biological effect of radiation is concerned with the evaluation of energy absorbed in the tissues. It was aimed to analyse the depth dose characteristics of x-ray beams of diverse energies to enhance the quality of radiotherapy treatment planning. Depth dose characteristics of different energy photon beams in water have been analysed. Photon beam is attenuated by the medium and the transmitted beam with less intensity causes lesser absorbed dose as depth increases. Relative attenuation on certain points on the beam axis and certain percentage of doses on different depths for available energies has been investigated. Photon beam depth dose characteristics do not show identical attributes as interaction of x-ray with matter is mainly governed by beam quality. Attenuation and penetration parameters of photon show variation with dosimetric parameters like field size due to scattering and Source to Surface Distance due to inverse square law, but the major parameter in photon interactions is its energy. Detailed analysis of photon Depth Dose characteristics helps to select appropriate beam for radiotherapy treatment when variety of beam energies available. Evaluation of this type of characteristics will help to establish theoretical relationships between dosimetric parameters to confirm measured values of dosimetric quantities, and hence to increase accuracy in radiotherapy treatment. (author)

  19. Calculation of dose-rate conversion factors for external exposure to photons and electrons

    International Nuclear Information System (INIS)

    Kocher, D.C.

    1978-01-01

    Methods are presented for the calculation of dose-rate conversion factors for external exposure to photon and electron radiation from radioactive decay. A dose-rate conversion factor is defined as the dose-equivalent rate per unit radionuclide concentration. Exposure modes considered are immersion in contaminated air, immersion in contaminated water, and irradiation from a contaminated ground surface. For each radiation type and exposure mode, dose-rate conversion factors are derived for tissue-equivalent material at the body surface of an exposed individual. In addition, photon dose-rate conversion factors are estimated for 22 body organs. The calculations are based on the assumption that the exposure medium is infinite in extent and that the radionuclide concentration is uniform. The dose-rate conversion factors for immersion in contaminated air and water then follow from the requirement that all of the energy emitted in the radioactive decay is absorbed in the infinite medium. Dose-rate conversion factors for ground-surface exposure are calculated at a reference location above a smooth, infinite plane using the point-kernel integration method and known specific absorbed fractions for photons and electrons in air

  20. Photon spectrum and absorbed dose in brain tumor

    Energy Technology Data Exchange (ETDEWEB)

    Silva S, A. [General Electric Healthcare, Antonio Dovali Jaime 70, Torre A 3er. piso, Col. Santa Fe, 01210 Mexico D. F. (Mexico); Vega C, H. R. [Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas, Zac. (Mexico); Rivera M, T. [IPN, Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Av. Legaria No. 694, 11500 Mexico D. F. (Mexico)

    2015-10-15

    Using Monte Carlo methods a BOMAB phantom inside a treatment hall with a brain tumor nearby the pituitary gland was treated with photons produced by a Varian 6 MV linac. The photon spectrum and the absorbed dose were calculated in the tumor, pituitary gland and the head. The treatment beam was collimated to illuminate only the tumor volume; however photons were noticed in the gland. Photon fluence reaching the tumor is 78.1 times larger than the fluence in the pituitary gland, on the other hand the absorbed dose in the tumor is 188 times larger than the dose in the gland because photons that reach the pituitary gland are scattered, by the head and the tumor, through Compton effect. (Author)

  1. Photon spectrum and absorbed dose in brain tumor.

    Science.gov (United States)

    Vega-Carrillo, Hector Rene; Silva-Sanchez, Angeles; Rivera-Montalvo, Teodoro

    2016-11-01

    Using Monte Carlo methods a BOMAB phantom inside a treatment hall with a brain tumor nearby the pituitary gland was treated with photons produced by a Varian 6MV linac. The photon spectrum and the absorbed dose were calculated in the tumor, pituitary gland and the head. The treatment beam was collimated to illuminate only the tumor volume; however photons were noticed in the gland. Photon fluence reaching the tumor is and 15.7 times larger than the fluence in the pituitary gland, on the other hand the absorbed dose in the tumor is 37.1 times larger than the dose in the gland because photons that reach the pituitary gland are scattered, by the head and the tumor, through Compton effect. Copyright © 2016 Elsevier Ltd. All rights reserved.

  2. Photon spectrum and absorbed dose in brain tumor

    International Nuclear Information System (INIS)

    Silva S, A.; Vega C, H. R.; Rivera M, T.

    2015-10-01

    Using Monte Carlo methods a BOMAB phantom inside a treatment hall with a brain tumor nearby the pituitary gland was treated with photons produced by a Varian 6 MV linac. The photon spectrum and the absorbed dose were calculated in the tumor, pituitary gland and the head. The treatment beam was collimated to illuminate only the tumor volume; however photons were noticed in the gland. Photon fluence reaching the tumor is 78.1 times larger than the fluence in the pituitary gland, on the other hand the absorbed dose in the tumor is 188 times larger than the dose in the gland because photons that reach the pituitary gland are scattered, by the head and the tumor, through Compton effect. (Author)

  3. Reassessment of calculation of effective dose equivalent for the CRCN-CO Environmental Radiological Monitoring Program

    International Nuclear Information System (INIS)

    Carneiro, L.B.; Dourado, M.A.; Barbosa, R.C.

    2017-01-01

    To reassess the calculations of the effective dose equivalent to obtain data of dosimetry and the accomplishment of the analysis comparing the data of several techniques that record doses of radiation originating from the cosmogenic and terrestrial contributions that make up the so-called background radiation. the basic information to be obtained is the contribution of the difference between the terrestrial dose equivalents, even the lowest concentration of primordial radionuclides, and that of the dose equivalent, deduced from TLD readings. (author)

  4. Reassessment of calculation of effective dose equivalent for the CRCN-CO Environmental Radiological Monitoring Program

    Energy Technology Data Exchange (ETDEWEB)

    Carneiro, L.B.; Dourado, M.A.; Barbosa, R.C., E-mail: research.photonics@gmail.com [Centro Regional de Ciências Nucleares do Centro-Oeste (CRCN-CO/CNEN-GO), Abadia de Goiás, GO (Brazil)

    2017-07-01

    To reassess the calculations of the effective dose equivalent to obtain data of dosimetry and the accomplishment of the analysis comparing the data of several techniques that record doses of radiation originating from the cosmogenic and terrestrial contributions that make up the so-called background radiation. the basic information to be obtained is the contribution of the difference between the terrestrial dose equivalents, even the lowest concentration of primordial radionuclides, and that of the dose equivalent, deduced from TLD readings. (author)

  5. Electron dose rate and photon contamination in electron arc therapy

    International Nuclear Information System (INIS)

    Pla, M.; Podgorsak, E.B.; Pla, C.

    1989-01-01

    The electron dose rate at the depth of dose maximum dmax and the photon contamination are discussed as a function of several parameters of the rotational electron beam. A pseudoarc technique with an angular increment of 10 degrees and a constant number of monitor units per each stationary electron field was used in our experiments. The electron dose rate is defined as the electron dose at a given point in phantom divided by the number of monitor units given for any one stationary electron beam. For a given depth of isocenter di the electron dose rates at dmax are linearly dependent on the nominal field width w, while for a given w the dose rates are inversely proportional to di. The dose rates for rotational electron beams with different di are related through the inverse square law provided that the two beams have (di,w) combinations which give the same characteristic angle beta. The photon dose at the isocenter depends on the arc angle alpha, field width w, and isocenter depth di. For constant w and di the photon dose at isocenter is proportional to alpha, for constant alpha and w it is proportional to di, and for constant alpha and di it is inversely proportional to w. The w and di dependence implies that for the same alpha the photon dose at the isocenter is inversely proportional to the electron dose rate at dmax

  6. Interaction of 2-Gy Equivalent Dose and Margin Status in Perioperative High-Dose-Rate Brachytherapy

    International Nuclear Information System (INIS)

    Martinez-Monge, Rafael; Cambeiro, Mauricio; Moreno, Marta; Gaztanaga, Miren; San Julian, Mikel; Alcalde, Juan; Jurado, Matias

    2011-01-01

    Purpose: To determine patient, tumor, and treatment factors predictive of local control (LC) in a series of patients treated with either perioperative high-dose-rate brachytherapy (PHDRB) alone (Group 1) or with PHDRB combined with external-beam radiotherapy (EBRT) (Group 2). Patient and Methods: Patients (n = 312) enrolled in several PHDRB prospective Phase I-II studies conducted at the Clinica Universidad de Navarra were analyzed. Treatment with PHDRB alone, mainly because of prior irradiation, was used in 126 patients to total doses of 32 Gy/8 b.i.d. or 40 Gy/10 b.i.d. treatments after R0 or R1 resections. Treatment with PHDRB plus EBRT was used in 186 patients to total doses of 16 Gy/4 b.i.d. or 24 Gy/6 b.i.d. treatments after R0 or R1 resections along with 45 Gy of EBRT with or without concomitant chemotherapy. Results: No dose-margin interaction was observed in Group 1 patients. In Group 2 patients there was a significant interaction between margin status and 2-Gy equivalent (Eq2Gy) dose (p = 0.002): (1) patients with negative margins had 9-year LC of 95.7% at Eq2Gy = 62.9Gy; (2) patients with close margins of >1 mm had 9-year LC of 92.4% at Eq2Gy = 72.2Gy, and (3) patients with positive/close <1-mm margins had 9-year LC of 68.0% at Eq2Gy = 72.2Gy. Conclusions: Two-gray equivalent doses ≥70 Gy may compensate the effect of close margins ≥1 mm but do not counterbalance the detrimental effect of unfavorable (positive/close <1 mm) resection margins. No dose-margin interaction is observed in patients treated at lower Eq2Gy doses ≤50 Gy with PHDRB alone.

  7. The energy spectrum of 662 keV photons in a water equivalent phantom

    International Nuclear Information System (INIS)

    Akar Tarim, U.; Gurler, O.; Ozmutlu, E.N.; Yalcin, S.; Gundogdu, O.; Sharaf, J.M.; Bradley, D.A.

    2012-01-01

    Investigation is made on the energy spectrum of photons originating from interactions of 662 keV primary gamma-ray photons emitted by a point source positioned at the centre of a water equivalent solid phantom of dimensions 19 cm×19 cm×24 cm. Peaks resulting from total energy loss (photopeak) and multiple and back scattering have been observed using a 51 mm×51 mm NaI(Tl) detector; good agreement being found between the measured and simulated response functions. The energy spectrum of the gamma photons obtained through the Monte Carlo simulation reveals local maxima at about 100 keV and 210 keV, being also observed in the experimental response function. Such spectra can be used as a method of testing the water equivalence of solid phantom media before their use for dosimetry measurements. - Highlights: ► Peaks resulting from total energy loss (photopeak) and multiple and back scattering were observed. ► Energy distribution of γ-ray photons from a point source at the centre of a water equivalent solid phantom. ► The method can be applied to various detector geometries.

  8. 10 CFR 835.203 - Combining internal and external equivalent doses.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 4 2010-01-01 2010-01-01 false Combining internal and external equivalent doses. 835.203 Section 835.203 Energy DEPARTMENT OF ENERGY OCCUPATIONAL RADIATION PROTECTION Standards for Internal and External Exposure § 835.203 Combining internal and external equivalent doses. (a) The total effective dose...

  9. Equivalent dose, effective dose and risk assessment from cephalometric radiography to critical organs

    International Nuclear Information System (INIS)

    Kang, Seong Sook; Cho, Bon Hae; Kim, Hyun Ja

    1995-01-01

    In head and neck region, the critical organ and tissue doses were determined, and the risks were estimated from lateral, posteroanterial and basilar cephalometric radiography. For each cephalometric radiography, 31 TLDs were placed in selected sites (18 internal and 13 external sites) in a tissue-equivalent phantom and exposed, then read-out in the TLD reader. The following results were obtained; 1. From lateral cephalometric radiography, the highest effective dose recorded was that delivered to the salivary gland (3.6 μSv) and the next highest dose was that received by the bone marrow (3 μSv). 2. From posteroanterial cephalometric radiography, the highest effective dose recorded was that delivered to the salivary gland (2 μSv) and the next highest dose was that received by the bone marrow (1.8 μSv). 3. From basilar cephalometric radiography, the highest effective dose recorded was that delivered to the thyroid gland (31.4 μSv) and the next highest dose was that received by the salivary gland (13.3 μSv). 4. The probabilities of stochastic effect from lateral, posteroanterial and basilar cephalometric radiography were 0.72 X 10 -6 , 0.49 X 10 -6 and 3.51 X 10 -6 , respectively.

  10. Dosimetric properties of a Solid Water High Equivalency (SW557) phantom for megavoltage photon beams.

    Science.gov (United States)

    Araki, Fujio

    2017-07-01

    The dosimetric properties of the recently developed SW557 phantom have been investigated by comparison with those of the existing SW457 phantom in megavoltage photon beams. The electron fluence ratio φ pl w , and chamber ionization ratio k pl , of water to SW457 and water to SW557 for 4-15MV photons were calculated as a function of depth using Monte Carlo simulations, and compared with measured values. Values of φ pl w for SW457 were in the range of 1.004-1.014 for 4MV, and 1.014-1.018 for 15MV photons. The φ pl w for SW557 ranged from 1.005 to 1.008 for 4MV and from 1.010 to 1.015 for 15MV photons and the variation of φ pl w with depth for each beam energy was within ±0.5%. Values of k pl were obtained with a PTW 30013 Farmer-type ionization chamber. The k pl for SW457 ranged from 0.997 to 1.011 for 4-15MV photons. Values of k pl for SW557 were almost unity for 4 and 6MV photons, while in the case of 10 and 15MV photons they were less than 1.006, excepting the build-up region. The measured and calculated k pl values of water to SW557 were in the range of 0.997-1.002 and 1.000-1.006, respectively, for 4-15MV photons, at a depth of 10cm with a source-to-axis distance of 100cm. The measured and calculated k pl values were in agreement within their uncertainty ranges. As a water-equivalent phantom, SW557 can be used with a dosimetric difference within±0.6%, for 4-15MV photons, and is more water-equivalent than SW457 in megavoltage photon beams. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  11. Method to account for dose fractionation in analysis of IMRT plans: Modified equivalent uniform dose

    International Nuclear Information System (INIS)

    Park, Clinton S.; Kim, Yongbok; Lee, Nancy; Bucci, Kara M.; Quivey, Jeanne M.; Verhey, Lynn J.; Xia Ping

    2005-01-01

    Purpose: To propose a modified equivalent uniform dose (mEUD) to account for dose fractionation using the biologically effective dose without losing the advantages of the generalized equivalent uniform dose (gEUD) and to report the calculated mEUD and gEUD in clinically used intensity-modulated radiotherapy (IMRT) plans. Methods and Materials: The proposed mEUD replaces the dose to each voxel in the gEUD formulation by a biologically effective dose with a normalization factor. We propose to use the term mEUD D o /n o that includes the total dose (D o ) and number of fractions (n o ) and to use the term mEUD o that includes the same total dose but a standard fraction size of 2 Gy. A total of 41 IMRT plans for patients with nasopharyngeal cancer treated at our institution between October 1997 and March 2002 were selected for the study. The gEUD and mEUD were calculated for the planning gross tumor volume (pGTV), planning clinical tumor volume (pCTV), parotid glands, and spinal cord. The prescription dose for these patients was 70 Gy to >95% of the pGTV and 59.4 Gy to >95% of the pCTV in 33 fractions. Results: The calculated average gEUD was 72.2 ± 2.4 Gy for the pGTV, 54.2 ± 7.1 Gy for the pCTV, 26.7 ± 4.2 Gy for the parotid glands, and 34.1 ± 6.8 Gy for the spinal cord. The calculated average mEUD D o /n o using 33 fractions was 71.7 ± 3.5 Gy for mEUD 70/33 of the pGTV, 49.9 ± 7.9 Gy for mEUD 59.5/33 of the pCTV, 27.6 ± 4.8 Gy for mEUD 26/33 of the parotid glands, and 32.7 ± 7.8 Gy for mEUD 45/33 of the spinal cord. Conclusion: The proposed mEUD, combining the gEUD with the biologically effective dose, preserves all advantages of the gEUD while reflecting the fractionation effects and linear and quadratic survival characteristics

  12. Assessment of organ equivalent doses and effective doses from diagnostic X-ray examinations

    International Nuclear Information System (INIS)

    Park, Sang Hyun

    2003-02-01

    The MIRD-type adult male, female and age 10 phantoms were constructed to evaluate organ equivalent dose and effective dose of patient due to typical diagnostic X-ray examination. These phantoms were constructed with external and internal dimensions of Korean. The X-ray energy spectra were generated with SPEC78. MCNP4B ,the general-purposed Monte Carlo code, was used. Information of chest PA , chest LAT, and abdomen AP diagnostic X-ray procedures was collected on the protocol of domestic hospitals. The results showed that patients pick up approximate 0.02 to 0.18 mSv of effective dose from a single chest PA examination, and 0.01 to 0.19 mSv from a chest LAT examination depending on the ages. From an abdomen AP examination, patients pick up 0.17 to 1.40 mSv of effective dose. Exposure time, organ depth from the entrance surface and X-ray beam field coverage considerably affect the resulting doses. Deviation among medical institutions is somewhat high, and this indicated that medical institutions should interchange their information and the need of education for medical staff. The methodology and the established system can be applied, with some expansion, to dose assessment for other medical procedures accompanying radiation exposure of patients like nuclear medicine or therapeutic radiology

  13. Measurement of the equivalent dose in quartz using a regenerative-dose single-aliquot protocol

    International Nuclear Information System (INIS)

    Murray, A.S.; Roberts, R.G.

    1998-01-01

    The principles behind a regenerative-dose single-aliquot protocol are outlined. It is shown for three laboratory-bleached Australian sedimentary quartz samples that the relative change in sensitivity of the optically stimulated luminescence (OSL) during a repeated measurement cycle (consisting of a dose followed by a 10 s preheat at a given temperature and then a 100 s exposure to blue/green light at 125 deg. C) is very similar to that of the 110 deg. C thermoluminescence (TL) peak measured during the preheat cycle. The absolute change in the TL sensitivity with preheat temperature is different for samples containing a natural or a regenerative dose. Furthermore, the absolute change in sensitivity in both the OSL and TL signals is non-linear with regeneration cycle, but the relative change in the OSL signal compared to the following 110 deg. C TL measurement is well approximated by a straight line. Both signals are thought to use the same luminescence centres, and so some common behaviour is not unexpected. A new regenerative-dose protocol is presented which makes use of this linear relationship to correct for sensitivity changes with regeneration cycle, and requires only one aliquot for the estimation of the equivalent dose (D e ). The protocol has been applied to quartz from nine Australian sites. To illustrate the value of the regenerative-dose single-aliquot approach, the apparent values of D e for 13 samples, containing doses of between 0.01 and 100 Gy, have been measured at various preheat temperatures of between 160 and 300 deg. C, using a single aliquot for each D e measurement. Excellent agreement is found between these single-aliquot estimates of D e and those obtained from additive-dose multiple-aliquot and single-aliquot protocols, over the entire dose range

  14. Dose conversion coefficients for photon exposure of the human eye lens

    Science.gov (United States)

    Behrens, R.; Dietze, G.

    2011-01-01

    In recent years, several papers dealing with the eye lens dose have been published, because epidemiological studies implied that the induction of cataracts occurs even at eye lens doses of less than 500 mGy. Different questions were addressed: Which personal dose equivalent quantity is appropriate for monitoring the dose to the eye lens? Is a new definition of the dose quantity Hp(3) based on a cylinder phantom to represent the human head necessary? Are current conversion coefficients from fluence to equivalent dose to the lens sufficiently accurate? To investigate the latter question, a realistic model of the eye including the inner structure of the lens was developed. Using this eye model, conversion coefficients for electrons have already been presented. In this paper, the same eye model—with the addition of the whole body—was used to calculate conversion coefficients from fluence (and air kerma) to equivalent dose to the lens for photon radiation from 5 keV to 10 MeV. Compared to the values adopted in 1996 by the International Commission on Radiological Protection (ICRP), the new values are similar between 40 keV and 1 MeV and lower by up to a factor of 5 and 7 for photon energies at about 10 keV and 10 MeV, respectively. Above 1 MeV, the new values (calculated without kerma approximation) should be applied in pure photon radiation fields, while the values adopted by the ICRP in 1996 (calculated with kerma approximation) should be applied in case a significant contribution from secondary electrons originating outside the body is present.

  15. Dose conversion coefficients for photon exposure of the human eye lens

    International Nuclear Information System (INIS)

    Behrens, R; Dietze, G

    2011-01-01

    In recent years, several papers dealing with the eye lens dose have been published, because epidemiological studies implied that the induction of cataracts occurs even at eye lens doses of less than 500 mGy. Different questions were addressed: Which personal dose equivalent quantity is appropriate for monitoring the dose to the eye lens? Is a new definition of the dose quantity H p (3) based on a cylinder phantom to represent the human head necessary? Are current conversion coefficients from fluence to equivalent dose to the lens sufficiently accurate? To investigate the latter question, a realistic model of the eye including the inner structure of the lens was developed. Using this eye model, conversion coefficients for electrons have already been presented. In this paper, the same eye model-with the addition of the whole body-was used to calculate conversion coefficients from fluence (and air kerma) to equivalent dose to the lens for photon radiation from 5 keV to 10 MeV. Compared to the values adopted in 1996 by the International Commission on Radiological Protection (ICRP), the new values are similar between 40 keV and 1 MeV and lower by up to a factor of 5 and 7 for photon energies at about 10 keV and 10 MeV, respectively. Above 1 MeV, the new values (calculated without kerma approximation) should be applied in pure photon radiation fields, while the values adopted by the ICRP in 1996 (calculated with kerma approximation) should be applied in case a significant contribution from secondary electrons originating outside the body is present.

  16. Changes in ambient dose equivalent rates around roads at Kawamata town after the Fukushima accident

    International Nuclear Information System (INIS)

    Kinase, Sakae; Sato, Satoshi; Yamamoto, Hideaki; Saito, Kimiaki; Sakamoto, Ryuichi

    2015-01-01

    Changes in ambient dose equivalent rates noted through vehicle-borne surveys have elucidated ecological half-lives of radioactive caesium in the environment. To confirm that the ecological half-lives are appropriate for predicting ambient dose equivalent rates within living areas, it is important to ascertain ambient dose equivalent rates on/around roads. In this study, radiation monitoring on/around roads at Kawamata town, located about 37 km northwest of the Fukushima Daiichi Nuclear Power Plant, was performed using monitoring vehicles and survey meters. It was found that the ambient dose equivalent rates around roads were higher than those on roads as of October 2012. And withal the ecological half-lives on roads were essentially consistent with those around roads. With dose predictions using ecological half-lives on roads, it is necessary to make corrections to ambient dose equivalent rates through the vehicle-borne surveys against those within living areas. (authors)

  17. Evaluation of two water-equivalent phantom materials for output calibration of photon and electron beams

    International Nuclear Information System (INIS)

    Liu Lizhong; Prasad, Satish C.; Bassano, Daniel A.

    2003-01-01

    Two commercially available water-equivalent solid phantom materials were evaluated for output calibration in both photon (6-15 MV) and electron (6-20 MeV) beams. The solid water 457 and virtual water materials have the same chemical composition but differ in manufacturing process and density. A Farmer-type ionization chamber was used for measuring the output of the photon beams at 5- and 10-cm depth and electron beams at maximum buildup depth in the solid phantoms and in natural water. The water-equivalency correction factor for the solid materials is defined as the ratio of the chamber reading in natural water to that in the solid at the same linear depth. For photon beams, the correction factor was found to be independent of depth and was 0.987 and 0.993 for 6- and 15-MV beams, respectively, for solid water. For virtual water, the corresponding correction factors were 0.993 and 0.998 for 6- and 15-MV beams, respectively. For electron beams, the correction factors ranged from 1.013 to 1.007 for energies of 6 to 20 MeV for both solid materials. This indicated that the water-equivalency of these materials is within ± 1.3%, making them suitable substitutes for natural water in both photon and electron beam output measurements over a wide energy range. These correction factors are slightly larger than the manufacturers' advertised values (± 1.0% for solid water and ± 0.5% for virtual water). We suggest that these corrections are large enough in most cases and should be applied in the calculation of beam outputs

  18. The development of BH3105E type neutron dose-equivalent meter

    International Nuclear Information System (INIS)

    Ji Changsong; Wang Tingting; Zhang Shuheng; Tan Baozeng

    2011-01-01

    A new BH3105E Type Neutron Dose-equivalent Meter has been developed. The 'multi-stick' ab- sorption method is used for thermal -14 MeV neutron equal dose-equivalent detection, what gives a high neutron sensitivity of 5 cps/μSv · h-1. RS-232 interface is accepted for signal communication (authors)

  19. Determination of Dose-Equivalent Response of A Typical Diamond Microdosimeter in Space Radiation Fields

    Directory of Open Access Journals (Sweden)

    firouz payervand

    2018-01-01

    Conclusion: The reasonable agreement between the dose equivalents calculated in this study and the results reported by other researchers confirmed that this type of microdosimeter could be a promising candidate suitable for the measurement of the dose equivalent in space radiation fields.

  20. Risk equivalent of exposure versus dose of radiation

    International Nuclear Information System (INIS)

    Bond, V.P.

    1986-01-01

    This report describes a risk analysis study of low-dose irradiation and the resulting biological effects on a cell. The author describes fundamental differences between the effects of high-level exposure (HLE) and low-level exposure (LLE). He stresses that the concept of absorbed dose to an organ is not a dose but a level of effect produced by a particular number of particles. He discusses the confusion between a linear-proportional representation of dose limits and a threshold-curvilinear representation, suggesting that a LLE is a composite of both systems

  1. Change of annual collective dose equivalent of radiation workers at KURRI

    International Nuclear Information System (INIS)

    Okamoto, Kenichi

    1994-01-01

    The change of exposure dose equivalent of radiation workers at KURRI (Kyoto University Research Reactor Institute) in the past 30 years is reported together with the operational accomplishments. The reactor achieved criticality on June 24, 1964 and reached the normal power of 1000 kW on August 17 of the same year, and the normal power was elevated to 5000 kW on July 16, 1968 until today. The change of the annual effective dose equivalent, the collective dose equivalent, the average annual dose equivalent and the maximum dose equivalent are indicated in the table and the figure. The chronological table on the activities of the reactor is added. (T.H.)

  2. Dose equivalent response of personal neutron dosemeters as a function of angle

    International Nuclear Information System (INIS)

    Tanner, J.E.; McDonald, J.C.; Stewart, R.D.; Wernli, C.

    1997-01-01

    The measured and calculated dose equivalent response as a function of angle has been examined for an albedo-type thermoluminescence dosemeter (TLD) that was exposed to unmoderated and D 2 O-moderated 252 Cf neutron sources while mounted on a 40 x 40 15 cm 3 polymethylmethacrylate phantom. The dosemeter used in this study is similar to many neutron personal dosemeters currently in use. The detailed construction of the dosemeter was modelled, and the dose equivalent response was calculated, using the MCNP code. Good agreement was found between the measured and calculated values of the relative dose equivalent angular response for the TLD albedo dosemeter. The relative dose equivalent angular response was also compared with the values of directional and personal dose equivalent as a function of angle published by Siebert and Schuhmacher. (author)

  3. Study of response of radiation monitors for environmental dose equivalent measurements

    International Nuclear Information System (INIS)

    Souza, Macilene N.; Khoury, H.J.

    2005-01-01

    The environmental dose equivalent H * (10), is the magnitude recommended by ICRU 39 for environmental monitoring in fields of radiation of photons. Most of the equipment used for area monitoring, only quantifies the magnitudes exposure or dose not being designed to this new magnitude. In Brazil, particularly, is not yet regulated the use of H * (10). However, with the revision of the standard 3.01 it will necessary the use of monitors that allow the achievement of measures according to H * (10). The transition for using new magnitudes will be a slow process and the contribution that the laboratories of metrology of ionizing radiation in the country can give is, at first, promote and create the habit of using the unit Sievert (Sv) in the calibration of the instruments, and that is the unit recommended for H * (10). In a second step, the tests for determining the response of the instruments for H * (10) should be made and this is the harder step, taking into account the large number of area monitors around the country. These tests will provide information about the limitations of the instrument to the new magnitude, that is, the range where the instrument will have the best performance in quantification of new magnitude. This paper evaluates the performance for H * (10), with the variation of energy and angle of incidence of radiation, of three of the most used monitors in the country

  4. Guideline values for skin decontamination measures based on nuclidspecific dose equivalent rate factors

    International Nuclear Information System (INIS)

    Pfob, H.; Heinemann, G.

    1992-01-01

    Corresponding dose equivalent rate factors for various radionuclides are now available for determining the skin dose caused by skin contamination. These dose equivalent rate factors take into account all contributions from the types of radiation emitted. Any limits for skin decontamination measures are nowhere contained or determined yet. However, radiological protection does in practice require at least guideline values in order to prevent unsuitable or detrimental measures that can be noticed quite often. New calculations of dose equivalent rate factors for the skin now make the recommendation of guideline values possible. (author)

  5. Two-photon equivalent weighting of spatial excimer laser beam profiles

    Science.gov (United States)

    Eva, Eric; Bauer, Harry H.; Metzger, K.; Pfeiffer, A.

    2001-04-01

    Damage in optical materials for semiconductor lithography applications caused by exposure to 248 or 193 nm light is usually two-photon driven, hence it is a nonlinear function of incident intensity. Materials should be tested with flat- topped temporal and spatial laser beam profiles to facilitate interpretation of data, but in reality this is hard to achieve. Sandstrom provided a formula that approximates any given temporal pulse shape with a two- photon equivalent rectangular pulse (Second Symposium on 193 nm Lithography, Colorado Springs 1997). Known as the integral-square pulse duration, this definition has been embraced as an industry standard. Originally faced with the problem of comparing results obtained with pseudo-Gaussian spatial profiles to literature data, we found that a general solution for arbitrarily inhomogeneous spatial beam profiles exists which results in a definition much similar to Sandstrom's. In addition, we proved the validity of our approach in experiments with intentionally altered beam profiles.

  6. Implementation of an Analytical Model for Leakage Neutron Equivalent Dose in a Proton Radiotherapy Planning System

    Energy Technology Data Exchange (ETDEWEB)

    Eley, John [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030 (United States); Graduate School of Biomedical Sciences, The University of Texas, 6767 Bertner Ave., Houston, TX 77030 (United States); Newhauser, Wayne, E-mail: newhauser@lsu.edu [Department of Physics and Astronomy, Louisiana State University and Agricultural and Mechanical College, 202 Nicholson Hall, Tower Drive, Baton Rouge, LA 70803 (United States); Mary Bird Perkins Cancer Center, 4950 Essen Lane, Baton Rouge, LA 70809 (United States); Homann, Kenneth; Howell, Rebecca [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030 (United States); Graduate School of Biomedical Sciences, The University of Texas, 6767 Bertner Ave., Houston, TX 77030 (United States); Schneider, Christopher [Department of Physics and Astronomy, Louisiana State University and Agricultural and Mechanical College, 202 Nicholson Hall, Tower Drive, Baton Rouge, LA 70803 (United States); Mary Bird Perkins Cancer Center, 4950 Essen Lane, Baton Rouge, LA 70809 (United States); Durante, Marco; Bert, Christoph [GSI Helmholtzzentrum für Schwerionenforschung, Planckstr. 1, Darmstadt 64291 (Germany)

    2015-03-11

    Equivalent dose from neutrons produced during proton radiotherapy increases the predicted risk of radiogenic late effects. However, out-of-field neutron dose is not taken into account by commercial proton radiotherapy treatment planning systems. The purpose of this study was to demonstrate the feasibility of implementing an analytical model to calculate leakage neutron equivalent dose in a treatment planning system. Passive scattering proton treatment plans were created for a water phantom and for a patient. For both the phantom and patient, the neutron equivalent doses were small but non-negligible and extended far beyond the therapeutic field. The time required for neutron equivalent dose calculation was 1.6 times longer than that required for proton dose calculation, with a total calculation time of less than 1 h on one processor for both treatment plans. Our results demonstrate that it is feasible to predict neutron equivalent dose distributions using an analytical dose algorithm for individual patients with irregular surfaces and internal tissue heterogeneities. Eventually, personalized estimates of neutron equivalent dose to organs far from the treatment field may guide clinicians to create treatment plans that reduce the risk of late effects.

  7. Fast neutron dose equivalent rates in heavy ion target areas

    International Nuclear Information System (INIS)

    Fulmer, C.B.; Butler, H.M.; Ohnesorge, W.F.; Mosko, S.W.

    1978-01-01

    At heavy ion accelerators, personnel access to areas near the target is sometimes important for successful performance of experiments. Radiation levels determine the amount of time that can be spent in these areas without exceeding maximum permissible exposures. Inasmuch as the fast neutrons contribute the major part of the Rem dose rates in these areas, knowledge of the fast neutron levels is important for planning permissive entry to target areas. Fast neutron dose rates were measured near thick medium mass targets bombarded with beams of C, N, O, and Ne ions. beam energies ranged from 3 to 16 MeV/amu. Dose rates (mrem/h) 1 meter from the target 90 degrees from the beam direction range from approx. 0.05 at MeV/amu to approx. 50 at 16 MeV/amu. These data should be helpful in planning permissive entry to heavy ion target areas

  8. Fast neutron dose equivalent rates in heavy ion target areas

    Energy Technology Data Exchange (ETDEWEB)

    Fulmer, C.B.; Butler, H.M.; Ohnesorge, W.F.; Mosko, S.W.

    1978-01-01

    At heavy ion accelerators, personnel access to areas near the target is sometimes important for successful performance of experiments. Radiation levels determine the amount of time that can be spent in these areas without exceeding maximum permissible exposures. Inasmuch as the fast neutrons contribute the major part of the Rem dose rates in these areas, knowledge of the fast neutron levels is important for planning permissive entry to target areas. Fast neutron dose rates were measured near thick medium mass targets bombarded with beams of C, N, O, and Ne ions. beam energies ranged from 3 to 16 MeV/amu. Dose rates (mrem/h) 1 meter from the target 90 degrees from the beam direction range from approx. 0.05 at MeV/amu to approx. 50 at 16 MeV/amu. These data should be helpful in planning permissive entry to heavy ion target areas.

  9. Variation of indoor radon concentration and ambient dose equivalent rate in different outdoor and indoor environments

    Energy Technology Data Exchange (ETDEWEB)

    Stojanovska, Zdenka; Janevik, Emilija; Taleski, Vaso [Goce Delcev University, Faculty of Medical Sciences, Stip (Macedonia, The Former Yugoslav Republic of); Boev, Blazo [Goce Delcev University, Faculty of Natural and Technical Sciences, Stip (Macedonia, The Former Yugoslav Republic of); Zunic, Zora S. [University of Belgrade, Institute of Nuclear Sciences ' ' Vinca' ' , Belgrade (Serbia); Ivanova, Kremena; Tsenova, Martina [National Center of Radiobiology and Radiation Protection, Sofia (Bulgaria); Ristova, Mimoza [University in Ss. Cyril and Methodius, Faculty of Natural Sciences and Mathematic, Institute of Physics, Skopje (Macedonia, The Former Yugoslav Republic of); Ajka, Sorsa [Croatian Geological Survey, Zagreb (Croatia); Bossew, Peter [German Federal Office for Radiation Protection, Berlin (Germany)

    2016-05-15

    Subject of this study is an investigation of the variations of indoor radon concentration and ambient dose equivalent rate in outdoor and indoor environments of 40 dwellings, 31 elementary schools and five kindergartens. The buildings are located in three municipalities of two, geologically different, areas of the Republic of Macedonia. Indoor radon concentrations were measured by nuclear track detectors, deployed in the most occupied room of the building, between June 2013 and May 2014. During the deploying campaign, indoor and outdoor ambient dose equivalent rates were measured simultaneously at the same location. It appeared that the measured values varied from 22 to 990 Bq/m{sup 3} for indoor radon concentrations, from 50 to 195 nSv/h for outdoor ambient dose equivalent rates, and from 38 to 184 nSv/h for indoor ambient dose equivalent rates. The geometric mean value of indoor to outdoor ambient dose equivalent rates was found to be 0.88, i.e. the outdoor ambient dose equivalent rates were on average higher than the indoor ambient dose equivalent rates. All measured can reasonably well be described by log-normal distributions. A detailed statistical analysis of factors which influence the measured quantities is reported. (orig.)

  10. Water-equivalent one-dimensional scintillating fiber-optic dosimeter for measuring therapeutic photon beam

    International Nuclear Information System (INIS)

    Moon, Jinsoo; Won Jang, Kyoung; Jae Yoo, Wook; Han, Ki-Tek; Park, Jang-Yeon; Lee, Bongsoo

    2012-01-01

    In this study, we fabricated a one-dimensional scintillating fiber-optic dosimeter, which consists of 9 scintillating fiber-optic dosimeters, septa, and PMMA blocks for measuring surface and percentage depth doses of a therapeutic photon beam. Each dosimeter embedded in the 1-D scintillating fiber-optic dosimeter is composed of square type organic scintillators and plastic optical fibers. Also black PVC films are used as septa to minimize cross-talk between the scintillating fiber-optic dosimeters. To construct a dosimeter system, a 1-D scintillating fiber-optic dosimeter and a CMOS image sensor were combined with 20 m-length plastic optical fibers. Using the dosimeter system, we measured surface and percentage depth doses of 6 and 15 MV photon beams and compared the results with those of EBT films and an ionization chamber. - Highlights: ► Fabrication of a one-dimensional scintillating fiber-optic dosimeter. ► The one-dimensional scintillating fiber-optic dosimeter has 9 scintillating fiber-optic dosimeters. ► Measurements of surface and percentage depth doses of a therapeutic photon beam. ► The results were compared with those of EBT films and an ionization chamber.

  11. Assessment of Effective Dose Equivalent of Indoor 222Rn Daughters in Inchass

    International Nuclear Information System (INIS)

    Ali, E.M.; Taha, T.M.; Gomaa, M.A.; El-Hussein, A.M.; Ahmad, A.A.

    2000-01-01

    The dominant component of natural radiation dose for the general population comes from the radon gas 222 Rn and its short-lived decay products, Ra A ( 214 Po), Ra B ( 214 Pb), Ra C ( 214 Bi), Ra C( 214 Po) in the breathing air. The objective of the present work is to assess the affective dose equivalent of the inhalation exposure of indoor 222 Rn for occupational workers. Average indon concentrations (Bqm -3 ) were monitored in several departments in Nuclear Research Center by radon monitor. We have calculated the lung dose equivalent and the effective dose equivalent for the Egyptian workers due to inhalation exposure of an equilibrium equivalent concentrations of radon daughters which varies from 0.27 to 2.5 mSvy -1 and 0.016 to 0.152mSvy -1 respectively. The annual effective doses obtained are within the accepted range of ICRP recommendations

  12. Evaluation of dose equivalent rate distribution in JCO critical accident by radiation transport calculation

    CERN Document Server

    Sakamoto, Y

    2002-01-01

    In the prevention of nuclear disaster, there needs the information on the dose equivalent rate distribution inside and outside the site, and energy spectra. The three dimensional radiation transport calculation code is a useful tool for the site specific detailed analysis with the consideration of facility structures. It is important in the prediction of individual doses in the future countermeasure that the reliability of the evaluation methods of dose equivalent rate distribution and energy spectra by using of Monte Carlo radiation transport calculation code, and the factors which influence the dose equivalent rate distribution outside the site are confirmed. The reliability of radiation transport calculation code and the influence factors of dose equivalent rate distribution were examined through the analyses of critical accident at JCO's uranium processing plant occurred on September 30, 1999. The radiation transport calculations including the burn-up calculations were done by using of the structural info...

  13. SU-F-T-408: On the Determination of Equivalent Squares for Rectangular Small MV Photon Fields

    International Nuclear Information System (INIS)

    Sauer, OA; Wegener, S; Exner, F

    2016-01-01

    Purpose: It is common practice to tabulate dosimetric data like output factors, scatter factors and detector signal correction factors for a set of square fields. In order to get the data for an arbitrary field, it is mapped to an equivalent square, having the same scatter as the field of interest. For rectangular fields both, tabulated data and empiric formula exist. We tested the applicability of such rules for very small fields. Methods: Using the Monte-Carlo method (EGSnrc-doseRZ), the dose to a point in 10cm depth in water was calculated for cylindrical impinging fluence distributions. Radii were from 0.5mm to 11.5mm with 1mm thickness of the rings. Different photon energies were investigated. With these data a matrix was constructed assigning the amount of dose to the field center to each matrix element. By summing up the elements belonging to a certain field, the dose for an arbitrary point in 10cm depth could be determined. This was done for rectangles up to 21mm side length. Comparing the dose to square field results, equivalent squares could be assigned. The results were compared to using the geometrical mean and the 4Xperimeter/area rule. Results: For side length differences less than 2mm, the difference between all methods was in general less than 0.2mm. For more elongated fields, relevant differences of more than 1mm and up to 3mm for the fields investigated occurred. The mean square side length calculated from both empiric formulas fitted much better, deviating hardly more than 1mm and for the very elongated fields only. Conclusion: For small rectangular photon fields, deviating only moderately from square both investigated empiric methods are sufficiently accurate. As the deviations often differ regarding their sign, using the mean improves the accuracy and the useable elongation range. For ratios larger than 2, Monte-Carlo generated data are recommended. SW is funded by Deutsche Forschungsgemeinschaft (SA481/10-1)

  14. SU-F-T-408: On the Determination of Equivalent Squares for Rectangular Small MV Photon Fields

    Energy Technology Data Exchange (ETDEWEB)

    Sauer, OA; Wegener, S; Exner, F [University of Wuerzburg, Wuerzburg (Germany)

    2016-06-15

    Purpose: It is common practice to tabulate dosimetric data like output factors, scatter factors and detector signal correction factors for a set of square fields. In order to get the data for an arbitrary field, it is mapped to an equivalent square, having the same scatter as the field of interest. For rectangular fields both, tabulated data and empiric formula exist. We tested the applicability of such rules for very small fields. Methods: Using the Monte-Carlo method (EGSnrc-doseRZ), the dose to a point in 10cm depth in water was calculated for cylindrical impinging fluence distributions. Radii were from 0.5mm to 11.5mm with 1mm thickness of the rings. Different photon energies were investigated. With these data a matrix was constructed assigning the amount of dose to the field center to each matrix element. By summing up the elements belonging to a certain field, the dose for an arbitrary point in 10cm depth could be determined. This was done for rectangles up to 21mm side length. Comparing the dose to square field results, equivalent squares could be assigned. The results were compared to using the geometrical mean and the 4Xperimeter/area rule. Results: For side length differences less than 2mm, the difference between all methods was in general less than 0.2mm. For more elongated fields, relevant differences of more than 1mm and up to 3mm for the fields investigated occurred. The mean square side length calculated from both empiric formulas fitted much better, deviating hardly more than 1mm and for the very elongated fields only. Conclusion: For small rectangular photon fields, deviating only moderately from square both investigated empiric methods are sufficiently accurate. As the deviations often differ regarding their sign, using the mean improves the accuracy and the useable elongation range. For ratios larger than 2, Monte-Carlo generated data are recommended. SW is funded by Deutsche Forschungsgemeinschaft (SA481/10-1)

  15. Skin dose estimation due to a contamination by a radionuclide β emitter: are doses equivalent good estimator of protection quantities?

    International Nuclear Information System (INIS)

    Bourgois, L.

    2011-01-01

    When handling radioactive β emitters, measurements in terms of personal dose equivalents H p (0.07) are used to estimate the equivalent dose limit to skin or extremities given by regulations. First of all, analytical expressions for individual dose equivalents H p (0.07) and equivalent doses to the extremities H skin are given for a point source and for contamination with a radionuclide β emitter. Second of all, operational quantities and protection quantities are compared. It is shown that in this case the operational quantities significantly overstate the protection quantities. For a skin contamination the ratio between operational quantities and protection quantities is 2 for a maximum β energy of 3 MeV and 90 for a maximum β energy of 150 keV. (author)

  16. Optimization of a neutron ambient dose equivalent rate meter

    International Nuclear Information System (INIS)

    Burgkhardt, B.; Fieg, G.; Piesch, E.; Klett, A.; Maushart, R.

    1994-01-01

    Co-operating in a technology transfer project, the Karlsruhe Nuclear Research Center and EG and G Berthold have developed a neutron equivalent doserate probe with high sensitivity and an energy dependent detection efficiency in accordance with the ICRP60 requirements. The special features of this probe were realized, on the one hand, by optimizing the moderator and absorber geometry through simulation calculations with the neutron transport code MCNP, and, on the other hand, by using a newly designed 3 He-methane proportional counter tube. The measurements were not yet completed when this paper went to press, however, it is to be expected that the response sensitivity will be more than 3 counts/nSv. (orig.) [de

  17. Criteria and methods for estimating external effective dose equivalent from personnel monitoring results: EDE implementation guide. Final report

    International Nuclear Information System (INIS)

    Owen, D.

    1998-09-01

    Title 10 Part 20 of the Code of Federal regulations requires that nuclear power plant licensees evaluate worker radiation exposure using a risk-based methodology termed the effective dose equivalent (EDE). EDE is a measure of radiation exposure that represents an individual's risk of stochastic injury from their exposure. EPRI has conducted research into how photons interact with the body. These results have been coupled with information on how the body's organs differ in their susceptibility to radiation injury, to produce a methodology for assessing the effective dose equivalent. The research and the resultant methodology have been described in numerous technical reports, scientific journal articles, and technical meetings. EPRI is working with the Nuclear Energy Institute to have the EPRI effective dose equivalent methodology accepted by the Nuclear Regulatory Commission for use at US nuclear power plants. In order to further familiarize power plant personnel with the methodology, this report summarizes the EDE research and presents some simple guidelines for its implementing the methodology

  18. Displacement damage analysis and modified electrical equivalent circuit for electron and photon-irradiated silicon solar cells

    Science.gov (United States)

    Arjhangmehr, Afshin; Feghhi, Seyed Amir Hossein

    2014-10-01

    Solar modules and arrays are the conventional energy resources of space satellites. Outside the earth's atmosphere, solar panels experience abnormal radiation environments and because of incident particles, photovoltaic (PV) parameters degrade. This article tries to analyze the electrical performance of electron and photon-irradiated mono-crystalline silicon (mono-Si) solar cells. PV cells are irradiated by mono-energetic electrons and poly-energetic photons and immediately characterized after the irradiation. The mean degradation of the maximum power (Pmax) of silicon solar cells is presented and correlated using the displacement damage dose (Dd) methodology. This method simplifies evaluation of cell performance in space radiation environments and produces a single characteristic curve for Pmax degradation. Furthermore, complete analysis of the results revealed that the open-circuit voltage (Voc) and the filling factor of mono-Si cells did not significantly change during the irradiation and were independent of the radiation type and fluence. Moreover, a new technique is developed that adapts the irradiation-induced effects in a single-cell equivalent electrical circuit and adjusts its elements. The "modified circuit" is capable of modeling the "radiation damage" in the electrical behavior of mono-Si solar cells and simplifies the designing of the compensation circuits.

  19. Application of combined TLD and CR-39 PNTD method for measurement of total dose and dose equivalent on ISS

    International Nuclear Information System (INIS)

    Benton, E.R.; Deme, S.; Apathy, I.

    2006-01-01

    To date, no single passive detector has been found that measures dose equivalent from ionizing radiation exposure in low-Earth orbit. We have developed the I.S.S. Passive Dosimetry System (P.D.S.), utilizing a combination of TLD in the form of the self-contained Pille TLD system and stacks of CR-39 plastic nuclear track detector (P.N.T.D.) oriented in three mutually orthogonal directions, to measure total dose and dose equivalent aboard the International Space Station (I.S.S.). The Pille TLD system, consisting on an on board reader and a large number of Ca 2 SO 4 :Dy TLD cells, is used to measure absorbed dose. The Pille TLD cells are read out and annealed by the I.S.S. crew on orbit, such that dose information for any time period or condition, e.g. for E.V.A. or following a solar particle event, is immediately available. Near-tissue equivalent CR-39 P.N.T.D. provides Let spectrum, dose, and dose equivalent from charged particles of LET ∞ H 2 O ≥ 10 keV/μm, including the secondaries produced in interactions with high-energy neutrons. Dose information from CR-39 P.N.T.D. is used to correct the absorbed dose component ≥ 10 keV/μm measured in TLD to obtain total dose. Dose equivalent from CR-39 P.N.T.D. is combined with the dose component <10 keV/μm measured in TLD to obtain total dose equivalent. Dose rates ranging from 165 to 250 μGy/day and dose equivalent rates ranging from 340 to 450 μSv/day were measured aboard I.S.S. during the Expedition 2 mission in 2001. Results from the P.D.S. are consistent with those from other passive detectors tested as part of the ground-based I.C.C.H.I.B.A.N. intercomparison of space radiation dosimeters. (authors)

  20. First results from electron-photon damage equivalence studies on a generic ethylene-propylene rubber

    International Nuclear Information System (INIS)

    Buckalew, W.H.

    1986-04-01

    As part of a simulator adequacy assessment program, the relative effectiveness of electrons and photons to produce damage in a generic ethylene propylene rubber (EPR) has been investigated. The investigation was limited in extent in that a single EPR material, in three thickness, was exposed to Cobalt-60 photons and three electron beam energies. Basing material damage on changes in the EPR mechanical properties elongation and tensile strength, we observed that EPR damage was a smoothly varying function of absorbed energy and independent of irradiating particle type. EPR damage tracked equally well as a function of both incident particle energy and material front surface dose. Based on these preliminary data, we tentatively concluded that a correlation between particle, particle energy, and material damage (as measured by changes in material elongation and/or tensile strength) has been demonstrated. 14 figs

  1. NEUTRON AND PHOTON DOSE MAPPING OF A DD NEUTRON GENERATOR.

    Science.gov (United States)

    Metwally, Walid A; Taqatqa, Osama A; Ballaith, Mohammed M; Chen, Allan X; Piestrup, Melvin A

    2017-11-01

    Neutron generators are an excellent tool that can be effectively utilized in educational institutions for applications such as neutron activation analysis, neutron radiography, and profiling and irradiation effects. For safety purposes, it is imperative that appropriate measures are taken in order to minimize the radiation dose from such devices to the operators, students and the public. This work presents the simulation and measurement results for the neutron and photon dose rates in the vicinity of the neutron generator installed at the University of Sharjah. A very good agreement is found between the simulated and measured dose rates. All of the public dose constraints were found to be met. The occupational dose constraint was also met after imposing a 200 cm no entry zone around the generator room. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  2. Risk equivalent of exposure versus dose of radiation

    International Nuclear Information System (INIS)

    Bond, V.P.

    1986-01-01

    Radiation is perhaps unique among all agents of interest in the Health Sciences in that it alone is both a therapeutic agent for the control of cancer and an essentially ubiquitous environmental agent with a potential for increasing the cancer rate in human populations. Therapy of tumors is accomplished with the high-level exposure (HLE) to radiation in order to effect control or a cure. Thus, it conforms to the concepts and approaches of pharmacology, toxicology, and therapeutic medicine. Only one function, that which relates the object-oriented and nonstochastic independent variable organ dose to its effect on a cancer or an organ, is needed to estimate the probability, P 2 , of a quantal response. Only P 2 is needed because P 1 , that the cancer slated for such treatment will receive some amount of the agent and be affected to some degree, is effectively unity. The health problem involving low-level exposure (LLE) to radiation, in contrast, is not at all analogous to those of pharmacology, toxicology, and medicine. Rather, it presents a public health problem in that it is a health population, albeit of cells, that is exposed in a radiation field composed of moving radiation particles with some attendant low-order carcinogenic or mutagenic risk. Thus, the concepts, quantities, and terminology applied to low-level radiation must be modified from their present orientation toward pharmacology, toxicology, medicine, and dose to conform to those of public health and accident statistics, in which both P 1 and P 2 for the exposed cells must be estimated

  3. Validity of the equivalent-photon approximation for the production of massive spin-1 particles

    International Nuclear Information System (INIS)

    Jayaraman, T.

    1986-01-01

    It is pointed out that the equivalent-photon approximation (EPA) for processes with massive spin-1 particles in the final state would have validity in a more restricted kinematic domain than for processes where it is commonly applied, viz., those with spin-1/2 or spin-0 particles in the final state. The criterion for the validity of EPA for the two-photon production of a pair of charged, massive, point-like spin-1 particles V ± , each of mass M and with a standard magnetic moment (k=1) is obtained. In a process in which one of the photons is real and the other virtual with four-momentum q, the condition for the validity of EPA is absolute value of q 2 2 , in addition to the usual condition absolute value of q 2 2 , W being the V + V - invariant mass. In a process in which both photons are virtual (with four-momenta q and q'), the condition is absolute value of q 2 absolute value of q' 2 W 4 8 , in addition to absolute value of q 2 2 , absolute value of q' 2 2 and absolute value of q 2 2 , absolute value of q' 2 2 . Even when these extra conditions permitting the use of EPA are not fulfilled, convenient approximate expressions may still be obtained assuming merely absolute value of q 2 2 and absolute value of q' 2 2 . It is also discussed how the extra conditions are altered when the vector bosons are incorporated in a spontaneously broken gauge theory. Examples of W boson production in Weinberg-Salam model are considered for which the condition absolute value of q 2 absolute value of q' 2 W 4 8 is shown to be removed. (author)

  4. Calculation of neutron fluence to dose equivalent conversion coefficients using GEANT4; Calculo de coeficientes de fluencia de neutrons para equivalente de dose individual utilizando o GEANT4

    Energy Technology Data Exchange (ETDEWEB)

    Ribeiro, Rosane M.; Santos, Denison de S.; Queiroz Filho, Pedro P. de; Mauricio, CLaudia L.P.; Silva, Livia K. da; Pessanha, Paula R., E-mail: rosanemribeiro@oi.com.br [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil)

    2014-07-01

    Fluence to dose equivalent conversion coefficients provide the basis for the calculation of area and personal monitors. Recently, the ICRP has started a revision of these coefficients, including new Monte Carlo codes for benchmarking. So far, little information is available about neutron transport below 10 MeV in tissue-equivalent (TE) material performed with Monte Carlo GEANT4 code. The objective of this work is to calculate neutron fluence to personal dose equivalent conversion coefficients, H{sub p} (10)/Φ, with GEANT4 code. The incidence of monoenergetic neutrons was simulated as an expanded and aligned field, with energies ranging between thermal neutrons to 10 MeV on the ICRU slab of dimension 30 x 30 x 15 cm{sup 3}, composed of 76.2% of oxygen, 10.1% of hydrogen, 11.1% of carbon and 2.6% of nitrogen. For all incident energy, a cylindrical sensitive volume is placed at a depth of 10 mm, in the largest surface of the slab (30 x 30 cm{sup 2}). Physic process are included for neutrons, photons and charged particles, and calculations are made for neutrons and secondary particles which reach the sensitive volume. Results obtained are thus compared with values published in ICRP 74. Neutron fluence in the sensitive volume was calculated for benchmarking. The Monte Carlo GEANT4 code was found to be appropriate to calculate neutron doses at energies below 10 MeV correctly. (author)

  5. Dose rate constants for new dose quantities

    International Nuclear Information System (INIS)

    Tschurlovits, M.; Daverda, G.; Leitner, A.

    1992-01-01

    Conceptual changes and new quantities made is necessary to reassess dose rate quantities. Calculations of the dose rate constant were done for air kerma, ambient dose equivalent and directional dose equivalent. The number of radionuclides is more than 200. The threshold energy is selected as 20 keV for the dose equivalent constants. The dose rate constant for the photon equivalent dose as used mainly in German speaking countries as a temporary quantity is also included. (Author)

  6. Neutron and photon dose assessment in Indus accelerator complex

    International Nuclear Information System (INIS)

    Verma, Dimple; Haridas Nair, G.; Bandopadhyay, Tapas; Tripathy, R.M.; Pal, Rupali; Bakshi, A.K.; Palani Selvam, T.; Datta, D.

    2016-02-01

    Indus Accelerator Complex (IAC) consists of 20 MeV Microtron, 450/550 MeV Booster, 450 MeV Indus-1 and 2.5 GeV Indus-2 storage rings. The radiation environment in Indus Accelerator Complex comprises of bremsstrahlung photons, electrons, positrons, photo neutrons and muons, out of which, bremsstrahlung photons are the major constituent of the prompt radiation. Major problem faced for on-line detection of neutrons is their severely pulsed nature. In the present study, measurement of neutron and photon dose rates in Indus Accelerator Complex was carried out using passive dosimeters such as CR-39 solid state nuclear track detector (SSNTD) and CaSO 4 :Dy Teflon disc, 6 LiF:Mg,Ti (TLD 600) and 7 LiF:Mg,Ti (TLD 700) based thermo luminescent (TL) detectors. The report describes the details of the measurement and discusses the results. (author)

  7. Longterm monitoring of ambient dose equivalent rates at aviation altitudes

    Energy Technology Data Exchange (ETDEWEB)

    Heber, B.; Burmeister, S.; Moeller, T.; Scharrenberg, E. [Institut fuer Experimentelle und Angewandte Physik, Christian-Albrechts-Universitaet zu Kiel, Kiel (Germany); Briese, J. [Deutsche Lufthansa AG, Frankfurt am Main (Germany); Burda, O.; Klages, T.; Langner, F.; Marquardt, J.; Wissmann, F. [Physikalisch-Technische Bundesanstalt, Bundesallee 100, 38116 Braunschweig Germany (Germany); Matthiae, D.; Reitz, G. [German Aerospace Center, Institute of Aerospace Medicine, Linder Hoehe, 51147 Koeln (Germany)

    2012-07-01

    The complex radiation field at flight altitudes results mainly from the interaction of energetic charged particles with atmospheric molecules and atoms and consists of secondary neutrons, protons, gamma rays, electrons, positrons and muons. Due to the continuous interactions of primary and secondary particles within the atmosphere, the intensity of each component depends on the height. Since the Earth's magnetic field acts as rigidity filter for the charged primary particles, the flux of the primary particles into the atmosphere and the resulting intensity of secondary particles depend on the geomagnetic latitude being highest over the geomagnetic poles. The main primary component consists of Galactic Cosmic Rays (GCRs), mainly protons and alpha particles, whose flux is modulated in the heliosphere. Beside this slowly varying galactic component, solar energetic particle events may temporarily increase the intensity of this radiation field. In the frame of the Radiation Monitoring on Board Aircraft (RAMONA) collaboration, three NAVIgation and DOSimetry (NAVIDOS) systems were installed in 2008 and 2009 on board of three Lufthansa Airbus A340 aircraft. They have been maintained since then by the consortium. Two of the NAVIDOS units rely on the DOSimetry TELescopes (DOSTELs), one is based on a LIULIN detector. This unique setup is ideally suited to investigate variations in the radiation field at different flight altitudes and geomagnetic positions and has been used to measure the radiation exposure during the recent extended solar minimum and thereafter. With increasing solar activity in 2010 the measured dose rates have been decreasing. Since these variations depend on the location of the aircraft, a detailed data analysis is required and presented.

  8. Tomographic anthropomorphic models. Pt. 4. Organ doses for adults due to idealized external photon exposures

    CERN Document Server

    Zankl, M; Petoussi-Henss, N; Regulla, D

    2002-01-01

    The present report contains extensive tables and figures of conversion coefficients of organ and tissue equivalent dose, normalised to air kerma free in air for voxel anthropomorphic phantoms and for standard geometries of external photon radiation, estimated with Monte Carlo techniques. Four realistic adult voxel phantoms were used for the calculations, based on computed tomographic data of real people: three male phantoms, two of them being of average size, one representing a big man, and one female phantom of a tall and somewhat over weighted woman.

  9. Tomographic anthropomorphic models. Pt. 4. Organ doses for adults due to idealized external photon exposures

    International Nuclear Information System (INIS)

    Zankl, M.; Petoussi-Henss, N.; Fill, U.; Regulla, D.

    2002-01-01

    The present report contains extensive tables and figures of conversion coefficients of organ and tissue equivalent dose, normalised to air kerma free in air for voxel anthropomorphic phantoms and for standard geometries of external photon radiation, estimated with Monte Carlo techniques. Four realistic adult voxel phantoms were used for the calculations, based on computed tomographic data of real people: three male phantoms, two of them being of average size, one representing a big man, and one female phantom of a tall and somewhat over weighted woman. (orig.)

  10. Applicability of ambient dose equivalent H*(d) in mixed radiation fields - a critical discussion

    International Nuclear Information System (INIS)

    Hajek, M.; Vana, N.

    2004-01-01

    For purposes of routine radiation protection, it is desirable to characterize the potential irradiation of individuals in terms of a single dose equivalent quantity that would exist in a phantom approximating the human body. The phantom of choice is the ICRU sphere made of 30 cm diameter tissue-equivalent plastic with a density of 1 g.cm-3 and a mass composition of 76.2 % O, 11.1 % C, 10.1 % H and 2.6 % N. Ambient dose equivalent, H*(d), was defined in ICRU report 51 as the dose equivalent that would be produced by an expanded and aligned radiation field at a depth d in the ICRU sphere. The recommended reference depths are 10 mm for strongly penetrating radiation and 0.07 mm for weakly penetrating radiation, respectively. As an operational quantity in radiation protection, H*(d) shall serve as a conservative and directly measurable estimate of protection quantities, e.g. effective dose E, which in turn are intended to give an indication of the risk associated with radiation exposure. The situation attains increased complexity in radiation environments being composed of a variety of charged and uncharged particles in a broad energetic spectrum. Radiation fields of similarly complex nature are, for example, encountered onboard aircraft and in space. Dose equivalent was assessed as a function of depth in quasi tissue-equivalent spheres by means of thermoluminescent dosemeters evaluated according to the high-temperature ratio (HTR) method. The presented experiments were performed both onboard aircraft and the Russian space station Mir. As a result of interaction processes within the phantom body, the incident primary spectrum may be significantly modified with increasing depth. For the radiation field at aviation altitudes we found the maximum of dose equivalent in a depth of 60 mm which conflicts with the 10 mm value recommended by ICRU. Contrary, for the space radiation environment the maximum dose equivalent was found at the surface of the sphere. This suggests that

  11. Applicability of Ambient Dose Equivalent H (d) in Mixed Radiation Fields - A Critical Discussion

    International Nuclear Information System (INIS)

    Vana, R.; Hajek, M.; Bergerm, T.

    2004-01-01

    For purposes of routine radiation protection, it is desirable to characterize the potential irradiation of individuals in terms of a single dose equivalent quantity that would exist in a phantom approximating the human body. The phantom of choice is the ICRU sphere made of 30 cm diameter tissue-equivalent plastic with a density of 1 g/cm3 and a mass composition of 76.2% O, 11.1% C, 10.1% H and 2.6% N. Ambient dose equivalent, H(d), was defined in ICRU report 51 as the dose equivalent that would be produced by an expanded and aligned radiation field at a depth d in the ICRU sphere. The recommended reference depths are 10 mm for strongly penetrating radiation and 0.07 mm for weakly penetrating radiation, respectively. As an operational quantity in radiation protection, H(d) shall serve as a conservative and directly measurable estimate of protection quantities, e.g. effective dose E, which in turn are intended to give an indication of the risk associated with radiation exposure. The situation attains increased complexity in radiation environments being composed of a variety of charged and uncharged particles in a broad energetic spectrum. Radiation fields of similarly complex nature are, for example, encountered onboard aircraft and in space. Dose equivalent was assessed as a function of depth in quasi tissue-equivalent spheres by means of thermoluminescent dosemeters evaluated according to the high-temperature ratio (HTR) method. The presented experiments were performed both onboard aircraft and the Russian space station Mir. As a result of interaction processes within the phantom body, the incident primary spectrum may be significantly modified with increasing depth. For the radiation field at aviation altitudes we found the maximum of dose equivalent in a depth of 60 mm which conflicts with the 10 mm value recommended by ICRU. Contrary, for the space radiation environment the maximum dose equivalent was found at the surface of the sphere. This suggests that skin

  12. Quality factor and dose equivalent investigations aboard the Soviet Space Station Mir

    Science.gov (United States)

    Bouisset, P.; Nguyen, V. D.; Parmentier, N.; Akatov, Ia. A.; Arkhangel'Skii, V. V.; Vorozhtsov, A. S.; Petrov, V. M.; Kovalev, E. E.; Siegrist, M.

    1992-07-01

    Since Dec 1988, date of the French-Soviet joint space mission 'ARAGATZ', the CIRCE device, had recorded dose equivalent and quality factor values inside the Mir station (380-410 km, 51.5 deg). After the initial gas filling two years ago, the low pressure tissue equivalent proportional counter is still in good working conditions. Some results of three periods are presented. The average dose equivalent rates measured are respectively 0.6, 0.8 and 0.6 mSv/day with a quality factor equal to 1.9. Some detailed measurements show the increasing of the dose equivalent rates through the SAA and near polar horns. The real time determination of the quality factors allows to point out high linear energy transfer events with quality factors in the range 10-20.

  13. Dose-rate conversion factors for external exposure to photon and electron radiation from radionuclides occurring in routine releases from nuclear fuel cycle facilities

    International Nuclear Information System (INIS)

    Kocher, D.C.

    1980-01-01

    Dose-rate conversion factors for external exposure to photon and electron radiation are calculated for 240 radionuclides of potential importance in routine releases from nuclear fuel cycle facilities. Exposure modes considered are immersion in contaminated air, immersion in contaminated water, and irradiation from a contaminated ground surface. For each exposure mode, dose-rate conversion factors for photons and electrons are calculated for tissue-equivalent material at the body surface of an exposed individual. Dose-rate conversion factors for photons only are calculated for 22 body organs. (author)

  14. Benefits of the effective dose equivalent concept at a medical center

    International Nuclear Information System (INIS)

    Vetter, R.J.; Classic, K.L.

    1991-01-01

    A primary objective of the recommendations of the International Committee on Radiological Protection Publication 26 is to insure that no source of radiation exposure is unjustified in relation to its benefits. This objective is consistent with goals of the Radiation Safety Committee and Institutional Review Board at medical centers where research may involve radiation exposure of human subjects. The effective dose equivalent concept facilitates evaluation of risk by those who have little or no knowledge of quantities or biological effects of radiation. This paper presents effective dose equivalent data used by radiation workers and those who evaluate human research protocols as these data relate to personal dosimeter reading, entrance skin exposure, and target organ dose. The benefits of using effective dose equivalent to evaluate risk of medical radiation environments and research protocols are also described

  15. Preliminary study of dose equivalent evaluation for residents in radioactivity contaminated rebar buildings

    International Nuclear Information System (INIS)

    Chen, W.L.; Liao, C.C.; Wang, M.T.; Chen, F. D.

    1998-01-01

    It has recently been found that several resident and office buildings in Taiwan were constructed with 60 Co-contaminated reinforcing steel bar (rebar). Both governmental officials and the residents of such buildings have been concerned about this finding. In order to respond to the situation, the government has adopted a number of remedial measures, including full-scale radiation survey, dose evaluation and physical examinations of residents. This article presents three methods for evaluating the dose equivalents of the residents living in the contaminated rebar buildings by means of γ-ray survey, necklace-type thermoluminescence dosimeters (TLDs) and the human lymphocyte chromosome aberration analyses. The results reveal that the dose evaluation by γ-ray survey is rather conservative. Generally for the residents whose annual dose equivalents are greater than 5 mSv (0.5 rem) by γ-ray survey, the dose equivalents from necklace-type TLDs are only within the range of 20 to 50% of the evaluated values mentioned above. For chromosome analyses, at least 500 lymphocyte cells were scored and analyzed for each resident. Most of the chromosome analysis data show that the dose equivalents received by residents are lower than the detection limit of the method (100 mSv) and quite different from the estimated dose obtained from either γ-ray survey or necklace-type TLD measurements

  16. Absorbed dose determination in photon fields using the tandem method

    International Nuclear Information System (INIS)

    Marques Pachas, J.F.

    1999-01-01

    The purpose of this work is to develop an alternative method to determine the absorbed dose and effective energy of photons with unknown spectral distributions. It includes a 'tandem' system that consists of two thermoluminescent dosemeters with different energetic dependence. LiF: Mg, Ti, CaF 2 : Dy thermoluminescent dosemeters and a Harshaw 3500 reading system are employed. Dosemeters are characterized with 90 Sr- 90 Y, calibrated with the energy of 60 Co and irradiated with seven different qualities of x-ray beams, suggested by ANSI No. 13 and ISO 4037. The answers of each type of dosemeter are adjusted to a function that depends on the effective energy of photons. The adjustment is carried out by means of the Rosenbrock minimization algorithm. The mathematical model used for this function includes five parameters and has a gauss and a straight line. Results show that the analytical functions reproduce the experimental data of the answers, with a margin of error of less than 5%. The reason of the answers of the CaF 2 : Dy and LiF: Mg, Ti, according to the energy of the radiation, allows us to establish the effective energy of photons and the absorbed dose, with a margin of error of less than 10% and 20% respectively

  17. The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Pt. 7. Organ doses due to parallel and environmental exposure geometries

    Energy Technology Data Exchange (ETDEWEB)

    Zankl, M. [GSF - Forschungszentrum fuer Umwelt und Gesundheit Neuherberg GmbH, Oberschleissheim (Germany). Inst. fuer Strahlenschutz; Drexler, G. [GSF - Forschungszentrum fuer Umwelt und Gesundheit Neuherberg GmbH, Oberschleissheim (Germany). Inst. fuer Strahlenschutz; Petoussi-Henss, N. [GSF - Forschungszentrum fuer Umwelt und Gesundheit Neuherberg GmbH, Oberschleissheim (Germany). Inst. fuer Strahlenschutz; Saito, K. [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan)

    1997-03-01

    This report presents a tabulation of organ and tissue equivalent dose as well as effective dose conversion coefficients, normalised to air kerma free in air, for occupational exposures and environmental exposures of the public to external photon radiation. For occupational exposures, whole-body irradiation with idealised geometries, i.e. broad parallel beams and fully isotropic radiation incidence, is considered. The directions of incidence for the parallel beams are anterior-posterior, posterior-anterior, left lateral, right lateral and a full 360 rotation around the body`s longitudinal axis. The influence of beam divergence on the body doses is also considered as well as the dependence of effective dose on the angle of radiation incidence. Regarding exposure of the public to environmental sources, three source geometries are considered: exposure from a radioactive cloud, from ground contamination and from the natural radionuclides distributed homogeneously in the ground. The precise angular and energy distributions of the gamma rays incident on the human body were taken into account. The organ dose conversion coefficients given in this catalogue were calculated using a Monte Carlo code simulating the photon transport in mathematical models of an adult male and an adult female, respectively. Conversion coefficients are given for the equivalent dose of 23 organs and tissues as well as for effective dose and the equivalent dose of the so-called `remainder`. The organ equivalent dose conversion coefficients are given separately for the adult male and female models and - as arithmetic mean of the conversion coefficients of both - for an average adult. Fitted data of the coefficients are presented in tables; the primary raw data as resulting from the Monte Carlo calculation are shown in figures together with the fitted data. (orig.)

  18. The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Pt. 7. Organ doses due to parallel and environmental exposure geometries

    International Nuclear Information System (INIS)

    Zankl, M.

    1997-03-01

    This report presents a tabulation of organ and tissue equivalent dose as well as effective dose conversion coefficients, normalised to air kerma free in air, for occupational exposures and environmental exposures of the public to external photon radiation. For occupational exposures, whole-body irradiation with idealised geometries, i.e. broad parallel beams and fully isotropic radiation incidence, is considered. The directions of incidence for the parallel beams are anterior-posterior, posterior-anterior, left lateral, right lateral and a full 360 rotation around the body's longitudinal axis. The influence of beam divergence on the body doses is also considered as well as the dependence of effective dose on the angle of radiation incidence. Regarding exposure of the public to environmental sources, three source geometries are considered: exposure from a radioactive cloud, from ground contamination and from the natural radionuclides distributed homogeneously in the ground. The precise angular and energy distributions of the gamma rays incident on the human body were taken into account. The organ dose conversion coefficients given in this catalogue were calculated using a Monte Carlo code simulating the photon transport in mathematical models of an adult male and an adult female, respectively. Conversion coefficients are given for the equivalent dose of 23 organs and tissues as well as for effective dose and the equivalent dose of the so-called 'remainder'. The organ equivalent dose conversion coefficients are given separately for the adult male and female models and - as arithmetic mean of the conversion coefficients of both - for an average adult. Fitted data of the coefficients are presented in tables; the primary raw data as resulting from the Monte Carlo calculation are shown in figures together with the fitted data. (orig.)

  19. Validation of SEACAB Methodology with Frascati (FNG) Photon Dose Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Tore, C.; Ortego, P.; Rodriguez Rivada, A.

    2014-07-01

    In the operation of the International Thermonuclear Experimental Reactor (ITER) the correct estimation of the gamma dose rate produced from the structural materials after shut down is one of the important safety parameter for hands-on maintenance. SEACAB, a rigorous 2-step (R2S) computational method has been developed for the calculation of residual dose in 3-D geometry with the use of the MCNP5 and of the ACAB (ACtivation ABacus) inventory code. The method is very efficient in hardware requirements being essentially modular. Starting from a single MCNP5 run permits a progressive improvement in the spatial detail of the material layers for the activation calculation and obtains separated source distributions for the isotopes contributing to the photon dose. (Author)

  20. Estimation of equivalent dose on the ends of hemodynamic physicians during neurological procedures

    International Nuclear Information System (INIS)

    Squair, Peterson L.; Souza, Luiz C. de; Oliveira, Paulo Marcio C. de

    2005-01-01

    The estimation of doses in the hands of physicists during hemodynamic procedures is important to verify the application of radiation protection related to the optimization and limit of dose, principles required by the Portaria 453/98 of Ministry of Health/ANVISA, Brazil. It was checked the levels of exposure of the hands of doctors during the use of the equipment in hemodynamic neurological procedures through dosimetric rings with thermoluminescent dosemeters detectors of LiF: Mg, Ti (TLD-100), calibrated in personal Dose equivalent HP (0.07). The average equivalent dose in the end obtained was 41.12. μSv per scan with an expanded uncertainty of 20% for k = 2. This value is relative to the hemodynamic Neurology procedure using radiological protection procedures accessible to minimize the dose

  1. Effectance, committed effective dose equivalent and annual limits on intake: what are the changes?

    International Nuclear Information System (INIS)

    Kendall, G.M.; Stather, J.W.; Phipps, A.W.

    1990-01-01

    This paper outlines the concept of effectance, compares committed effectance with the old committed effective dose equivalent and goes on to discuss changes in the annual limits on intakes and the maximum organ doses which would result from an intake of an ALI (Annual Limit of Intake). It is shown that committed effectance is usually, but not always, higher than committed effective dose equivalent. ALIS are usually well below those resulting from the ICRP Publication 30 scheme. However, if the ALI were based only on a limit on effectance it would imply a high dose to specific organs for certain nuclides. In order to control maximum organ doses an explicit limit could be introduced. However, this would destroy some of the attractive features of the new scheme. An alternative would be a slight modification to some of the weighting factors. (author)

  2. Metoprolol Dose Equivalence in Adult Men and Women Based on Gender Differences: Pharmacokinetic Modeling and Simulations

    Directory of Open Access Journals (Sweden)

    Andy R. Eugene

    2016-11-01

    Full Text Available Recent meta-analyses and publications over the past 15 years have provided evidence showing there are considerable gender differences in the pharmacokinetics of metoprolol. Throughout this time, there have not been any research articles proposing a gender stratified dose-adjustment resulting in an equivalent total drug exposure. Metoprolol pharmacokinetic data was obtained from a previous publication. Data was modeled using nonlinear mixed effect modeling using the MONOLIX software package to quantify metoprolol concentration–time data. Gender-stratified dosing simulations were conducted to identify equivalent total drug exposure based on a 100 mg dose in adults. Based on the pharmacokinetic modeling and simulations, a 50 mg dose in adult women provides an approximately similar metoprolol drug exposure to a 100 mg dose in adult men.

  3. Quantifying the Combined Effect of Radiation Therapy and Hyperthermia in Terms of Equivalent Dose Distributions

    International Nuclear Information System (INIS)

    Kok, H. Petra; Crezee, Johannes; Franken, Nicolaas A.P.; Stalpers, Lukas J.A.; Barendsen, Gerrit W.; Bel, Arjan

    2014-01-01

    Purpose: To develop a method to quantify the therapeutic effect of radiosensitization by hyperthermia; to this end, a numerical method was proposed to convert radiation therapy dose distributions with hyperthermia to equivalent dose distributions without hyperthermia. Methods and Materials: Clinical intensity modulated radiation therapy plans were created for 15 prostate cancer cases. To simulate a clinically relevant heterogeneous temperature distribution, hyperthermia treatment planning was performed for heating with the AMC-8 system. The temperature-dependent parameters α (Gy −1 ) and β (Gy −2 ) of the linear–quadratic model for prostate cancer were estimated from the literature. No thermal enhancement was assumed for normal tissue. The intensity modulated radiation therapy plans and temperature distributions were exported to our in-house-developed radiation therapy treatment planning system, APlan, and equivalent dose distributions without hyperthermia were calculated voxel by voxel using the linear–quadratic model. Results: The planned average tumor temperatures T90, T50, and T10 in the planning target volume were 40.5°C, 41.6°C, and 42.4°C, respectively. The planned minimum, mean, and maximum radiation therapy doses were 62.9 Gy, 76.0 Gy, and 81.0 Gy, respectively. Adding hyperthermia yielded an equivalent dose distribution with an extended 95% isodose level. The equivalent minimum, mean, and maximum doses reflecting the radiosensitization by hyperthermia were 70.3 Gy, 86.3 Gy, and 93.6 Gy, respectively, for a linear increase of α with temperature. This can be considered similar to a dose escalation with a substantial increase in tumor control probability for high-risk prostate carcinoma. Conclusion: A model to quantify the effect of combined radiation therapy and hyperthermia in terms of equivalent dose distributions was presented. This model is particularly instructive to estimate the potential effects of interaction from different treatment

  4. Comparison of different dose calculation methods for irregular photon fields

    International Nuclear Information System (INIS)

    Zakaria, G.A.; Schuette, W.

    2000-01-01

    In this work, 4 calculation methods (Wrede method, Clarskon method of sector integration, beam-zone method of Quast and pencil-beam method of Ahnesjoe) are introduced to calculate point doses in different irregular photon fields. The calculations cover a typical mantle field, an inverted Y-field and different blocked fields for 4 and 10 MV photon energies. The results are compared to those of measurements in a water phantom. The Clarkson and the pencil-beam method have been proved to be the methods of equal standard in relation to accuracy. Both of these methods are being distinguished by minimum deviations and applied in our clinical routine work. The Wrede and beam-zone methods deliver useful results to central beam and yet provide larger deviations in calculating points beyond the central axis. (orig.) [de

  5. EFFDOS - a FORTRAN-77-code for the calculation of the effective dose equivalent

    International Nuclear Information System (INIS)

    Baer, M.; Honcu, S.; Huebschmann, W.

    1984-01-01

    The FORTRAN-77-code EFFDOS calculates the effective dose equivalent according to ICRP 26 due to the longterm emission of radionuclides into the atmosphere for the following exposure pathways: inhalation, ingestion, γ-ground irradiation (γ-irradiation by radionuclides deposited on the ground) and β- or γ-submersion (irradiation by the passing radioactive cloud). For calculating the effective dose equivalent at a single spot it is necessary to put in the diffusion factor and - if need be - the washout factor; otherwise EFFDOS calculates the input data for the computer codes ISOLA III and WOLGA-1, which then are enabled to compute the atmospheric diffusion, ground deposition and local dose equivalent distribution for the requested exposure pathway. Atmospheric diffusion, deposition and radionuclide transfer are calculated according to the ''Allgemeine Berechnungsgrundlage ....'' recommended by the German Fed. Ministry of Interior. A sample calculated is added. (orig.) [de

  6. Evaluation of dose equivalent to the people accompanying patients in diagnostic radiology using MCNP4C Monte Carlo code

    International Nuclear Information System (INIS)

    Mehdizadeh, S.; Faghihi, R.; Sina, S.; Zehtabian, M.

    2007-01-01

    Complete text of publication follows. Objective: X rays used in diagnostic radiology contribute a major share to population doses from man-made sources of radiation. In some branches of radiology, it is necessary that another person stay in the imaging room and immobilize the patient to carry out radiological operation. ICRP 70 recommends that this should be done by parents or accompanying nursing or ancillary personnel and not in any case by radiation workers. Methods: Dose measurements were made previously using standard methods employing LiF TLD-100 dosimeters. A TLD card was installed on the main trunk of the body of the accompanying people where the maximum dose was probable. In this research the general purpose Monte Carlo N-particle radiation transport computer code (MCNP4C) is used to calculate the equivalent dose to the people accompanying patients exposed to radiation scattered from the patient (Without protective clothing). To do the simulations, all components of the geometry are placed within an air-filled box. Two homogeneous water phantoms are used to simulate the patient and the accompanying person. The accompanying person leans against the table at one side of the patient. Finally in case of source specification, only the focus of the X-ray tube is modelled, i.e. as a standard MCNP point source emitting a cone of photons. Photon stopping material is used as a collimator model to reduce the circular cross section of the cone to a rectangle. The X-ray spectra to be used in the MCNP simulations are generated with spectrum generator software, taking the X-ray voltage and all filtration applied in the clinic as input parameters. These calculations are done for different patient sizes and for different radiological operations. Results: In case of TL dosimetry, for a group of 100 examinations, the dose equivalents ranged from 0.01 μsv to 0.13 msv with the average of 0.05 msv. The results are seen to be in close agreement with Monte Carlo simulations

  7. MOSFET dosimeter depth-dose measurements in heterogeneous tissue-equivalent phantoms at diagnostic x-ray energies

    International Nuclear Information System (INIS)

    Jones, A.K.; Pazik, F.D.; Hintenlang, D.E.; Bolch, W.E.

    2005-01-01

    The objective of the present study was to explore the use of the TN-1002RD metal-oxide-semiconductor field effect transistor (MOSFET) dosimeter for measuring tissue depth dose at diagnostic photon energies in both homogeneous and heterogeneous tissue-equivalent materials. Three cylindrical phantoms were constructed and utilized as a prelude to more complex measurements within tomographic physical phantoms of pediatric patients. Each cylindrical phantom was constructed as a stack of seven 5-cm-diameter and 1-cm-thick discs of materials radiographically representative of either soft tissue (S), bone (B), or lung tissue (L) at diagnostic photon energies. In addition to a homogeneous phantom of soft tissue (SSSSSSS), two heterogeneous phantoms were constructed: SSBBSSS and SBLLBSS. MOSFET dosimeters were then positioned at the interface of each disc, and the phantoms were then irradiated at 66 kVp and 200 mAs. Measured values of absorbed dose at depth were then compared to predicated values of point tissue dose as determined via Monte Carlo radiation transport modeling. At depths exceeding 2 cm, experimental results matched the computed values of dose with high accuracy regardless of the dosimeter orientation (epoxy bubble facing toward or away from the x-ray beam). Discrepancies were noted, however, between measured and calculated point doses near the surface of the phantom (surface to 2 cm depth) when the dosimeters were oriented with the epoxy bubble facing the x-ray beam. These discrepancies were largely eliminated when the dosimeters were placed with the flat side facing the x-ray beam. It is therefore recommended that the MOSFET dosimeters be oriented with their flat sides facing the beam when they are used at shallow depths or on the surface of either phantoms or patients

  8. Influence of thermoluminescence trapping parameter from abundant quartz powder on equivalent dose

    International Nuclear Information System (INIS)

    Zhao Qiuyue; Wei Mingjian; Song Bo; Pan Baolin; Zhou Rui

    2014-01-01

    Glow curves of abundant quartz powder were obtained with the RGD-3B thermoluminescence (TL) reader. TL peaks with 448, 551, 654, 756 K were identified at the heating rate of 5 K/s. The activation energy, frequency factor and lifetime of trapped charge were evaluated at ambient temperature for four peaks by the method of various heating rates. Within a certain range of activation energy, the equivalent dose increases exponentially with the activation energy. The equivalent dose increases from 54 Gy to 485 Gy with the temperature from 548 K to 608 K, and it fluctuates around 531 Gy with the temperature from 608 K to 748 K. (authors)

  9. Neutron fluence-to-dose equivalent conversion factors: a comparison of data sets and interpolation methods

    International Nuclear Information System (INIS)

    Sims, C.S.; Killough, G.G.

    1983-01-01

    Various segments of the health physics community advocate the use of different sets of neutron fluence-to-dose equivalent conversion factors as a function of energy and different methods of interpolation between discrete points in those data sets. The major data sets and interpolation methods are used to calculate the spectrum average fluence-to-dose equivalent conversion factors for five spectra associated with the various shielded conditions of the Health Physics Research Reactor. The results obtained by use of the different data sets and interpolation methods are compared and discussed. (author)

  10. Prediction analysis of dose equivalent responses of neutron dosemeters used at a MOX fuel facility

    International Nuclear Information System (INIS)

    Tsujimura, N.; Yoshida, T.; Takada, C.

    2011-01-01

    To predict how accurately neutron dosemeters can measure the neutron dose equivalent (rate) in MOX fuel fabrication facility work environments, the dose equivalent responses of neutron dosemeters were calculated by the spectral folding method. The dosemeters selected included two types of personal dosemeter, namely a thermoluminescent albedo neutron dosemeter and an electronic neutron dosemeter, three moderator-based neutron survey meters, and one special instrument called an H p (10) monitor. The calculations revealed the energy dependences of the responses expected within the entire range of neutron spectral variations observed in neutron fields at workplaces. (authors)

  11. SU-E-T-43: Analytical Model for Photon Peripheral Dose in Radiotherapy Treatments

    Energy Technology Data Exchange (ETDEWEB)

    Nieto, B Sanchez; El far, R [Instituto de Fisica, Pontificia Universidad Catolica de Chile, Santiago, Santiago De Chile (Chile); Romero-Exposito, M [Universitat Autonoma de Barcelona, Barcelona (Spain); Lagares, J [Centro de Investigaciones Energeticas Medioambientales y Tecnologicas, Madrid (Spain); Mateo, JC [Hospital Duques del Infantado, Sevilla (Spain); Terron, JA [Servicio de Radiofisica, Hospital Universitario Virgen Macarena, Sevilla (Spain); Irazola, L; Sanchez-Doblado, F [Servicio de Radiofisica, Hospital Universitario Virgen Macarena, Sevilla (Spain); Departamento de Fisiologia Medica y Biofisica, Universidad de Sevilla, Sevilla (Spain)

    2014-06-01

    Purpose: The higher survival rate of radiotherapy patients entails a growing concern on second cancers associated to peripheral doses. Currently, dosimetry of out-of field doses is still under development. Our group has developed a methodology to estimate neutron equivalent dose in organs (1,2). We aimed to propose a model to estimate out-of-field photon doses in isocentric treatments from basic clinical data. Methods: The proposed function models the dose as the sum of leakage and scatter terms. The latter is modeled as a virtual source at the collimator, which suffers from attenuation in air and tissue, corrected by the inverse-square-law. The model was parameterized using experimental measurements with TLD700 chips placed inside an anthropomorphic phantom (6–18MV) irradiated with conformal and modulated techniques in Elekta, Siemens and Varian linacs. This model provides photon dose at a point as a function of clinical parameters as prescription dose/UM, PTV volume, distance to the field edge, height of the MLC leaves and distance from the the MLC to the isocenter. Model was tested against independent measurements (TLD100) for a VMAT treatment on a Elekta. Dose to organs is modeled from dose to points along the head-to-feet axis of the organ of a “standard man” escalated by patient height. Results: Our semi-empirical model depends on 3 given parameters (leakage parameter can be individualized). A novelty of our model, over other models (e.g., PERIDOSE), arises from its applicability to any technique (independently of the number of MU needed to deliver a dose). Differences between predictions and measurements were < 0.005mSv/UM. Conclusion: We have proposed a unique model which successfully account for photon peripheral organ dose. This model can be applied in the day-to-day clinic as it only needs a few basic parameters which are readily accessible.1. Radiother. Oncol. 107:234–243, 2013. 2. Phys. Med. Biol. 57:6167–6191, 2012.

  12. SU-E-T-43: Analytical Model for Photon Peripheral Dose in Radiotherapy Treatments

    International Nuclear Information System (INIS)

    Nieto, B Sanchez; El far, R; Romero-Exposito, M; Lagares, J; Mateo, JC; Terron, JA; Irazola, L; Sanchez-Doblado, F

    2014-01-01

    Purpose: The higher survival rate of radiotherapy patients entails a growing concern on second cancers associated to peripheral doses. Currently, dosimetry of out-of field doses is still under development. Our group has developed a methodology to estimate neutron equivalent dose in organs (1,2). We aimed to propose a model to estimate out-of-field photon doses in isocentric treatments from basic clinical data. Methods: The proposed function models the dose as the sum of leakage and scatter terms. The latter is modeled as a virtual source at the collimator, which suffers from attenuation in air and tissue, corrected by the inverse-square-law. The model was parameterized using experimental measurements with TLD700 chips placed inside an anthropomorphic phantom (6–18MV) irradiated with conformal and modulated techniques in Elekta, Siemens and Varian linacs. This model provides photon dose at a point as a function of clinical parameters as prescription dose/UM, PTV volume, distance to the field edge, height of the MLC leaves and distance from the the MLC to the isocenter. Model was tested against independent measurements (TLD100) for a VMAT treatment on a Elekta. Dose to organs is modeled from dose to points along the head-to-feet axis of the organ of a “standard man” escalated by patient height. Results: Our semi-empirical model depends on 3 given parameters (leakage parameter can be individualized). A novelty of our model, over other models (e.g., PERIDOSE), arises from its applicability to any technique (independently of the number of MU needed to deliver a dose). Differences between predictions and measurements were < 0.005mSv/UM. Conclusion: We have proposed a unique model which successfully account for photon peripheral organ dose. This model can be applied in the day-to-day clinic as it only needs a few basic parameters which are readily accessible.1. Radiother. Oncol. 107:234–243, 2013. 2. Phys. Med. Biol. 57:6167–6191, 2012

  13. Annual dose equivalents estimation received by Cienfuegos population due medical practice

    International Nuclear Information System (INIS)

    Usagaua R, Z.; Santander I, E.

    1996-01-01

    This study represents the first evaluation of the effective equivalent dose that receives the population of the Cienfuegos province in Cuba because of medical practice. The evaluation is based on the tables of doses depending on several parameters that influence over these ones, and also based on large diagnostic examinations statistics of all medical institutions over a 9 years period. Values of examinations frequency, contribution to total dose from radiography, fluoroscopy, dental radiography and nuclear medicine, and other characteristics of the last ones are offered. A comparative reflection dealing with received doses by radiography and fluoroscopy techniques is also included. (authors). 4 refs

  14. The neutron dose equivalent around high energy medical electron linear accelerators

    Directory of Open Access Journals (Sweden)

    Poje Marina

    2014-01-01

    Full Text Available The measurement of neutron dose equivalent was made in four dual energy linear accelerator rooms. Two of the rooms were reconstructed after decommissioning of 60Co units, so the main limitation was the space. The measurements were performed by a nuclear track etched detectors LR-115 associated with the converter (radiator that consist of 10B and with the active neutron detector Thermo BIOREM FHT 742. The detectors were set at several locations to evaluate the neutron ambient dose equivalent and/or neutron dose rate to which medical personnel could be exposed. Also, the neutron dose dependence on collimator aperture was analyzed. The obtained neutron dose rates outside the accelerator rooms were several times smaller than the neutron dose rates inside the accelerator rooms. Nevertheless, the measured neutron dose equivalent was not negligible from the aspect of the personal dosimetry with almost 2 mSv a year per person in the areas occupied by staff (conservative estimation. In rooms with 15 MV accelerators, the neutron exposure to the personnel was significantly lower than in the rooms having 18 MV accelerators installed. It was even more pronounced in the room reconstructed after the 60Co decommissioning. This study confirms that shielding from the neutron radiation should be considered when building vaults for high energy linear accelerators, especially when the space constraints exist.

  15. Absorbed dose measurement by using tissue equivalent ionization chamber (pair ionization chamber) in the Yayoi reactor

    International Nuclear Information System (INIS)

    Sasuga, N.; Okamura, K.; Terakado, T.; Mabuchi, Y.; Nakagawa, T.; Sukegawa, Toshio; Aizawa, C.; Saito, I.; Oka, Yoshiaki

    1998-01-01

    Each dose rate of neutron and gamma ray in the thermal column of the Yayoi reactor, in which an epithermal neutron field will be used for the boron neutron capture therapy, was measured by using a tissue equivalent ionization chamber and a graphite chamber. The tissue equivalent ionization chamber has some response to both neutron and gamma ray, but the graphite chamber has a few response to the neutron, so called pair ionization chamber method. The epithermal neutron fluxes of the thermal column were calculated by ANISN (one dimensional neutron-gamma transport code). A measured value for gamma dose rate by the pair ionization chamber agrees relevantly with a calculated result. For neutron dose rate, however, the measured value was too much small in comparison with the calculated result. The discrepancy between the measured value and the calculated result for neutron dose rate is discussed in detail in the report. (M. Suetake)

  16. Preliminary characterization of the passive neutron dose equivalent monitor with TLDs

    Energy Technology Data Exchange (ETDEWEB)

    Tsujimura, Norio; Kanai, Katsuta; Momose, Takumaro; Hayashi, Naomi [Japan Nuclear Cycle Development Inst., Tokai Works, Tokai, Ibaraki (Japan); Chen Erhu [Beijing Institute of Nuclear Engineering, Beijing (China)

    2001-02-01

    The passive neutron dose equivalent monitor with TLDs is composed of a cubic polyethylene moderator and TLDs at the center of moderator. This monitor was originally designed for measurements of neutron doses over long-term period of time around the nuclear facilities. In this study, the energy response of this monitor was calculated by Monte Carlo methods and experimentally obtained under {sup 241}Am-Be, {sup 252}Cf and moderated {sup 252}Cf neutron irradiation. Additionally, the responses of two types of conventional neutron dose equivalent meters (rem counters) were also investigated as comparison. The authors concluded that this passive neutron monitor with TLDs had a good energy response similar to conventional rem counters and could evaluate neutron doses within 10% of accuracy to the moderated fission spectra. (author)

  17. Dependence of alanine gel dosimeter response as a function of photon clinical beams dose rate

    International Nuclear Information System (INIS)

    Silva, Cleber Feijo; Campos, Leticia Lucente

    2013-01-01

    Gel dosimetry is a new area developed by Gore, it is ery useful for application in radiotherapy because using NMR imaging as evaluation technique is possible to evaluate three dimensional absorbed dose distribution. The measure technique is based on difference of ferrous (Fe 2+ ) and ferric (Fe 3+ ) ) ions concentration that can be measured also by spectrophotometry technique. The Alanine gel dosimeter was developed at IPEN. The alanine is an amino acid and tissue equivalent material that presents significant improvement on previous alanine dosimetry systems. The addition of Alanine increases the production of ferric ions in the solution. This work aims to study the dose rate dependence of photon clinical beams radiation on the alanine gel dosimeter optical response, as well as the response repeatability and gel production reproducibility, since this property is very important for characterization and standardization of any dosimeter. (author)

  18. Organ and effective dose coefficients for cranial and caudal irradiation geometries: photons

    International Nuclear Information System (INIS)

    Veinot, K.G.; Eckerman, K.F.; Hertel, N.E.

    2016-01-01

    With the introduction of new recommendations of the International Commission on Radiological Protection (ICRP) in Publication 103, the methodology for determining the protection quantity, effective dose, has been modified. The modifications include changes to the defined organs and tissues, the associated tissue weighting factors, radiation weighting factors and the introduction of reference sex-specific computational phantoms. Computations of equivalent doses in organs and tissues are now performed in both the male and female phantoms and the sex-averaged values used to determine the effective dose. Dose coefficients based on the ICRP 103 recommendations were reported in ICRP Publication 116, the revision of ICRP Publication 74 and ICRU Publication 57. The coefficients were determined for the following irradiation geometries: anterior-posterior (AP), posterior-anterior (PA), right and left lateral (RLAT and LLAT), rotational (ROT) and isotropic (ISO). In this work, the methodology of ICRP Publication 116 was used to compute dose coefficients for photon irradiation of the body with parallel beams directed upward from below the feet (caudal) and directed downward from above the head (cranial). These geometries may be encountered in the workplace from personnel standing on contaminated surfaces or volumes and from overhead sources. Calculations of organ and tissue kerma and absorbed doses for caudal and cranial exposures to photons ranging in energy from 10 keV to 10 GeV have been performed using the MCNP6.1 radiation transport code and the adult reference phantoms of ICRP Publication 110. As with calculations reported in ICRP 116, the effects of charged-particle transport are evident when compared with values obtained by using the kerma approximation. At lower energies the effective dose per particle fluence for cranial and caudal exposures is less than AP orientations while above ∼30 MeV the cranial and caudal values are greater. (authors)

  19. Parallel processing of dose calculation for external photon beam therapy

    International Nuclear Information System (INIS)

    Kunieda, Etsuo; Ando, Yutaka; Tsukamoto, Nobuhiro; Ito, Hisao; Kubo, Atsushi

    1994-01-01

    We implemented external photon beam dose calculation programs into a parallel processor system consisting of Transputers, 32-bit processors especially suitable for multi-processor configuration. Two network conformations, binary-tree and pipeline, were evaluated for rectangular and irregular field dose calculation algorithms. Although computation speed increased in proportion to the number of CPU, substantial overhead caused by inter-processor communication occurred when a smaller computation load was delivered to each processor. On the other hand, for irregular field calculation, which requires more computation capability for each calculation point, the communication overhead was still less even when more than 50 processors were involved. Real-time responses could be expected for more complex algorithms by increasing the number of processors. (author)

  20. A study on the annual equivalent doses received by cardiologists in a UK hospital

    International Nuclear Information System (INIS)

    Fong, R.Y.L.; Ryan, E.; Alonso-Arrizabalaga, S.

    2001-01-01

    A dose assessment study was carried out to determine the likely annual equivalent doses received by various parts of a cardiologist's body. High sensitivity GR-200 thermoluminescent dosemeters were attached to cardiologists' foreheads, little fingers, wrists, elbows, knees and ankles. Three common cardiology procedures were investigated, namely, percutaneous transluminal coronary angioplasty (PTCA), permanent pacemaker insertion (PPM) and left heart catheterisation (LHC). Dose monitoring was done on a case-by-case basis. Data on ten cases of each procedure were gathered. The projected annual equivalent doses were computed by averaging the ten doses measured at each site for each examination type and finding out from the cardiologists how many cases of PTCA, PPM and LHC they do in a year. Results in this study show that for the lens of the eye, the projected annual equivalent dose is below 10 mSv and for the other body parts, it is below 100 mSv per year. The study demonstrated that the methodology used can help to optimise radiation protection in diagnostic radiology. (author)

  1. Electron fluence to dose equivalent conversion factors calculated with EGS3 for electrons and positrons with energies from 100 keV to 20 GeV

    International Nuclear Information System (INIS)

    Rogers, D.W.O.

    1983-01-01

    At NRC the general purpose Monte-Carlo electron-photon transport code EGS3 is being applied to a variety of radiation dosimetry problems. To test its accuracy at low energies a detailed set of depth-dose curves for electrons and photons has been generated and compared to previous calculations. It was found that by changing the default step-size algorithm in EGS3, significant changes were obtained for incident electron beam cases. It was found that restricting the step-size to a 4% energy loss was appropriate below incident electron beam energies of 10 MeV. With this change, the calculated depth-dose curves were found to be in reasonable agreement with other calculations right down to incident electron energies of 100 keV although small (less than or equal to 10%) but persistent discrepancies with the NBS code ETRAN were obtained. EGS3 predicts higher initial dose and shorter range than ETRAN. These discrepancies are typical of a wide range of energies as is the better agreement with the results of Nahum. Data is presented for the electron fluence to maximal dose equivalent in a 30 cm thick slab of ICRU 4-element tissue irradiated by broad parallel beams of electrons incident normal to the surface. On their own, these values only give an indication of the dose equivalent expected from a spectrum of electrons since one needs to fold the spectrum maximal dose equivalent value. Calculations have also been done for incident positron beams. Despite the large statistical uncertainties, maximal dose equivalent although their values are 5 to 10% lower in a band around 10 MeV

  2. Additional effective dose equivalent for adults and children in Poland as the result of mushroom consumption

    International Nuclear Information System (INIS)

    Jasinska, H.; Kozak, K.; Mietelski, J.W.

    2004-01-01

    Experimental data of caesium radioactivity in samples of various mushrooms collected all over Poland from 1986 to 1989 are presented. Nearly 80 samples from Poland and a few samples from Austria and USSR were analysed. The effective dose equivalents for adults and children caused by the consumption of one mass unit of dried mushrooms for each sample were estimated. (author)

  3. Evaluation of the effective dose equivalent to tbe public of Pavia after the Chernobyl nuclear accident

    International Nuclear Information System (INIS)

    Altieri, S.; Berzero, A.; Meloni, S.; Rosti, G.; Genova, N.

    1988-01-01

    The Chernobyl radionuclide monitoring campaign in air particulate and foodstuffs was carried out and continued up to June 1987. On the basis of collected data estimates of the collective effective dose equivalent commitment to the public of province of Pavia, by external irradiation or by inhalation, were carried out and are reported in the present paper

  4. Dependence on age at intake of committed dose equivalents from radionuclides

    International Nuclear Information System (INIS)

    Adams, N.

    1981-01-01

    The dependence of committed dose equivalents on age at intake is needed to assess the significance of exposures of young persons among the general public resulting from inhaled or ingested radionuclides. The committed dose equivalents, evaluated using ICRP principles, depend on the body dimensions of the young person at the time of intake of a radionuclide and on subsequent body growth. Representation of growth by a series of exponential segments facilitates the derivation of general expressions for the age dependence of committed dose equivalents if metabolic models do not change with age. The additional assumption that intakes of radionuclides in air or food are proportional to a person's energy expenditure (implying age-independent dietary composition) enables the demonstration that the age of the most highly exposed 'critical groups' of the general public from these radionuclides is either about 1 year or 17 years. With the above assumptions the exposure of the critical group is less than three times the exposure of adult members of the general public. Approximate values of committed dose equivalents which avoid both underestimation and excessive overestimation are shown to be obtainable by simplified procedures. Modified procedures are suggested for use if metabolic models change with age. (author)

  5. Radiobiological equivalent of low/high dose rate brachytherapy and evaluation of tumor and normal responses to the dose.

    Science.gov (United States)

    Manimaran, S

    2007-06-01

    The aim of this study was to compare the biological equivalent of low-dose-rate (LDR) and high-dose-rate (HDR) brachytherapy in terms of the more recent linear quadratic (LQ) model, which leads to theoretical estimation of biological equivalence. One of the key features of the LQ model is that it allows a more systematic radiobiological comparison between different types of treatment because the main parameters alpha/beta and micro are tissue-specific. Such comparisons also allow assessment of the likely change in the therapeutic ratio when switching between LDR and HDR treatments. The main application of LQ methodology, which focuses on by increasing the availability of remote afterloading units, has been to design fractionated HDR treatments that can replace existing LDR techniques. In this study, with LDR treatments (39 Gy in 48 h) equivalent to 11 fractions of HDR irradiation at the experimental level, there are increasing reports of reproducible animal models that may be used to investigate the biological basis of brachytherapy and to help confirm theoretical predictions. This is a timely development owing to the nonavailability of sufficient retrospective patient data analysis. It appears that HDR brachytherapy is likely to be a viable alternative to LDR only if it is delivered without a prohibitively large number of fractions (e.g., fewer than 11). With increased scientific understanding and technological capability, the prospect of a dose equivalent to HDR brachytherapy will allow greater utilization of the concepts discussed in this article.

  6. The Effect of Breast Reconstruction Prosthesis on Photon Dose Distribution in Breast Cancer Radiotherapy

    Directory of Open Access Journals (Sweden)

    fatemeh sari

    2017-12-01

    Full Text Available Introduction: Siliconeprosthetic implants are commonlyutilizedfor tissue replacement and breast augmentation after mastectomy. On the other hand, some patients require adjuvant radiotherapy in order to preventlocal-regional recurrence and increment ofthe overall survival. In case of recurrence, the radiation oncologist might have to irradiate the prosthesis.The aim of this study was to evaluate the effect of silicone prosthesis on photon dose distribution in breast radiotherapy. Materials and Methods: The experimental dosimetry was performed using theprosthetic breast phantom and the female-equivalent mathematical chest phantom. A Computerized Tomographybased treatment planning was performedusing a phantom and by CorePlan Treatment Planning System (TPS. For measuring the absorbed dose, thermoluminescent dosimeter(TLD chips (GR-207A were used. Multiple irradiations were completed for all the TLD positions, and the dose absorbed by the TLDs was read by a lighttelemetry (LTM reader. Results: Statistical comparisons were performed between the absorbed dosesassessed by the TLDs and the TPS calculations forthe same sites. Our initial resultsdemonstratedanacceptable agreement (P=0.064 between the treatment planning data and the measurements. The mean difference between the TPS and TLD resultswas 1.99%.The obtained findings showed that radiotherapy is compatible withsilicone gel prosthesis. Conclusion: It could be concludedthat the siliconbreast prosthesis has no clinicallysignificant effectondistribution of a 6 MV photon beam for reconstructed breasts.

  7. Mathematical modelling for dose deposition in photon-therapy

    International Nuclear Information System (INIS)

    Pichard, Teddy

    2016-01-01

    Radiotherapy treatments consists in irradiating the patient with beams of energetic particles (typically photons) targeting the tumor. Such particles are transported through the medium and deposit energy in the medium. This deposited energy is the so called dose, responsible for the biological effect of the radiations. The present work aim to develop numerical methods for dose computation and optimization that are competitive in terms of computational cost and accuracy compared to reference method. The motion of particles is first studied through a system of linear transport equations at the kinetic level. However, solving directly such systems is numerically too costly for medical application. Instead, the moment method is used with a special focus on the Mn models. Those moment equations are non-linear and valid under a condition called realizability. Standard numerical schemes for moment equations are constrained by stability conditions which happen to be very restrictive when the medium contains low density regions. Inconditionally stable numerical schemes adapted to moment equations (preserving the realizability property) are developed. Those schemes are shown to be competitive in terms of computational costs compared to reference approaches. Finally they are applied to in an optimization procedure aiming to maximize the dose in the tumor and to minimize the dose in healthy tissues. (author) [fr

  8. Estimates of effective equivalent dose commitments for Slovene population following the Chernobyl accident

    International Nuclear Information System (INIS)

    Kanduc, M.; Jovanowic, O.; Kuhar, B.

    2004-01-01

    This paper shows the estimates of effective equivalent dose commitments for the two groups of Slovene population, 5 years old children and adults. Doses were calculated on the basis of the ICRP 30 methodology, first from the measurements of the concentrations of the radionuclides in air, water and food samples and then compared with the results of the measurements of radionuclides in composite samples of the prepared food, taken in the kindergarten nearby. Results show that there is certain degree of conservatism hidden in the calculation of the doses on the basis of measurements of the activity concentration in the elements of the biosphere and is estimated to be roughly 50%. (author)

  9. Application of the personnel photographic monitoring method to determine equivalent radiation dose beyond proton accelerator shielding

    International Nuclear Information System (INIS)

    Gel'fand, E.K.; Komochkov, M.M.; Man'ko, B.V.; Salatskaya, M.I.; Sychev, B.S.

    1980-01-01

    Calculations of regularities to form radiation dose beyond proton accelerator shielding are carried out. Numerical data on photographic monitoring dosemeter in radiation fields investigated are obtained. It was shown how to determine the total equivalent dose of radiation fields beyond proton accelerator shielding by means of the photographic monitoring method by introduction into the procedure of considering nuclear emulsions of division of particle tracks into the black and grey ones. A comparison of experimental and calculational data has shown the applicability of the used calculation method for modelling dose radiation characteristics beyond proton accelerator shielding [ru

  10. Distribution of dose within the body from a photon emitter present in an organ

    International Nuclear Information System (INIS)

    Snyder, W.S.; Ford, M.R.; Warner, G.G.

    1977-01-01

    A dosimetric system was developed which provides estimates of mean radiation dose to organs from photon sources distributed uniformly in one or more organs. Although the sources of photons are assumed to be distributed uniformly, it is not true that dose from these photons is uniformly distributed. In particular, when a source of photons is located in a particular organ, nearby tissues will be irradiated at doses which decrease markedly with distance from the source. The mean dose may give a poor approximation to the actual dose if the tissues over which dose is averaged are extensive, for example, the remainder of the body. A set of enveloping organs was devised for liver, lungs, etc., which give mean dose at distances from zero to one centimeter from the source organ, from one to two centimeters, etc. These can be used to yield estimates of the extent of inhomogeneity of the dose distribution from a source of photons located in the source organ

  11. Equivalent dose, effective dose and risk assessment from panoramic radiography to the critical organs of head and neck region

    International Nuclear Information System (INIS)

    Cho, Bong Hae; Nah, Kyung Soo; Lee, Ae Ryeon

    1995-01-01

    The purpose of this study was to evaluate the equivalent and effective dose, and estimate radiation risk to the critical organs of head and neck region from the use of adult and child mode in panoramic radiography. The results were as follows. 1. The salivary glands showed the highest equivalent and effective dose in adult and child mode. The equivalent and effective dose in adult mode were 837 μSv and 20.93 μSv, those in child mode were 462 μSv and 11.54 μSv, respectively. 2. Total effective doses to the critical head and neck organs were estimated 34.2l μSv in adult mode, 20.14 μSv in child mode. From these data, the probabilities of stochastic effect from adult and child mode were 2.50xl0 -6 and 1.47x10 -6 3. The other remainder showed the greatest risk of fatal cancer. The risk estimate were 4.5 and 2.7 fatal malignancies in adult and child mode from million examinations. The bone marrow and thyroid gland showed about 0.1 fatal cancer in adult. and child mode from these examinations.

  12. The Effect of Aquaplast on Surface Dose of Photon Beam

    International Nuclear Information System (INIS)

    Oh, Do Hoon; Bae, Hoon Sik

    1995-01-01

    Purpose : To evaluate the effect on surface dose due to Aquaplast used for immobilizing the patients with head and neck cancers in photon beam radiotherapy. Materials and Methods : To assess surface and buildup region dose for 6MV X-ray from linear accelerator(Siemens Mevatron 6740), we measured percent ionization value with the Markus chamber model 30-329 manufactured by PTW Frieburg and Capintec electrometer, model WK92. For measurement of surface ionization value, the chamber was embedded in 25 X 25 X 3 cm 3 acrylic phantom and set on 25 X 25 X 5 cm 3 , polystyrene phantom to allow adequate scattering. The measurements of percent depth ionization were made by placing the polystyrene layers of appropriate thickness over the chamber. The measurements were taken at 10 cm SSD for 5 X 5 cm 2 , 10 X 10 cm 2 , and 15 X 15 cm 2 field sizes, respectively. Placing the layer of Aquaplast over the chamber, the same procedures were repeated. We evaluated two types o Aquaplast: 1.6mm layer of original Aquaplast(manufactured by WFR Aquaplast Corp.) and transformed Aquaplast similar to moulded one for immobilizing the patients practically. We also measured surface ionization values with blocking tray in presence or absence of transformed Aquaplast. In calculating percent depth dose, we used the formula suggested by Gerbi and khan to correct over response of the Markus chamber. Results : The surface doses for open fields of 5 X 5 cm 2 , 10 X 10 cm 2 , 15 X 15 cm 2 were 7.9%, 13.6%, and 18.7% respectively. He original Aquaplast increased the surface doses upto 38.4%, 43.6% and 47.4% respectively. There were little differences in percent depth dose values beyond the depth of Dmax. Increasing field size, the blocking tray caused increase of the surface dose by 0.2%, 1.7%, 3.0% without Aquaplast, 0.2%, 1.9%, 3.7% with transformed Aquaplast, respectively. Conclusion : The original and transformed Aquaplast increased the surface dose moderately. The percent depth doses beyond Dmax

  13. Method for calculating individual equivalent doses and cumulative dose of population in the vicinity of nuclear power plant site

    International Nuclear Information System (INIS)

    Namestek, L.; Khorvat, D; Shvets, J.; Kunz, Eh.

    1976-01-01

    A method of calculating the doses of external and internal person irradiation in the nuclear power plant vicinity under conditions of normal operation and accident situations has been described. The main difference between the above method and methods used up to now is the use of a new antropomorphous representation of a human body model together with all the organs. The antropomorphous model of human body and its organs is determined as a set of simple solids, coordinates of disposistion of the solids, sizes, masses, densities and composition corresponding the genuine organs. The use of the Monte-Carlo method is the second difference. The results of the calculations according to the model suggested can be used for determination: a critical group of inhabitans under conditions of normal plant operation; groups of inhabitants most subjected to irradiation in the case of possible accident; a critical sector with a maximum collective dose in the case of an accident; a critical radioisotope favouring the greatest contribution to an individual equivalent dose; critical irradiation ways promoting a maximum contribution to individual equivalent doses; cumulative collective doses for the whole region or for a chosen part of the region permitting to estimate a population dose. The consequent method evoluation suggests the development of separate units of the calculationg program, critical application and the selection of input data of physical, plysiological and ecological character and improvement of the calculated program for the separate concrete events [ru

  14. Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium

    International Nuclear Information System (INIS)

    Carrasco, P.; Jornet, N.; Duch, M.A.; Weber, L.; Ginjaume, M.; Eudaldo, T.; Jurado, D.; Ruiz, A.; Ribas, M.

    2004-01-01

    An extensive set of benchmark measurement of PDDs and beam profiles was performed in a heterogeneous layer phantom, including a lung equivalent heterogeneity, by means of several detectors and compared against the predicted dose values by different calculation algorithms in two treatment planning systems. PDDs were measured with TLDs, plane parallel and cylindrical ionization chambers and beam profiles with films. Additionally, Monte Carlo simulations by meansof the PENELOPE code were performed. Four different field sizes (10x10, 5x5, 2x2, and1x1 cm 2 ) and two lung equivalent materials (CIRS, ρ e w =0.195 and St. Bartholomew Hospital, London, ρ e w =0.244-0.322) were studied. The performance of four correction-based algorithms and one based on convolution-superposition was analyzed. The correction-based algorithms were the Batho, the Modified Batho, and the Equivalent TAR implemented in the Cadplan (Varian) treatment planning system and the TMS Pencil Beam from the Helax-TMS (Nucletron) treatment planning system. The convolution-superposition algorithm was the Collapsed Cone implemented in the Helax-TMS. The only studied calculation methods that correlated successfully with the measured values with a 2% average inside all media were the Collapsed Cone and the Monte Carlo simulation. The biggest difference between the predicted and the delivered dose in the beam axis was found for the EqTAR algorithm inside the CIRS lung equivalent material in a 2x2 cm 2 18 MV x-ray beam. In these conditions, average and maximum difference against the TLD measurements were 32% and 39%, respectively. In the water equivalent part of the phantom every algorithm correctly predicted the dose (within 2%) everywhere except very close to the interfaces where differences up to 24% were found for 2x2 cm 2 18 MV photon beams. Consistent values were found between the reference detector (ionization chamber in water and TLD in lung) and Monte Carlo simulations, yielding minimal differences (0

  15. Dose Measurements of Bremsstrahlung-Produced Neutrons at the Advanced Photon Source

    International Nuclear Information System (INIS)

    Job, P.K.; Pisharody, M.; Semones, E.

    1998-01-01

    Bremsstrahlung is generated in the storage rings of the synchrotron radiation facilities by the radiative interaction of the circulating particle beam with both the residual gas molecules and storage ring components. These bremsstrahlung photons, having an energy range of zero to the maximum energy of the particle beam, interact with beamline components like beam stops and collimators generating photoneutrons of varying energies. There are three main processes by which photoneutrons may be produced by the high energy bremsstrahlung photons: giant nuclear dipole resonance and decay (10 MeV γ γ γ > 140 MeV). The giant resonance neutrons are emitted almost isotropically and have an average energy of about 2 MeV. High energy neutrons (E > 10 MeV) emitted from the quasi-deuteron decay and intranuclear cascade are peaked in the forward direction. At the Advanced Photon Source (APS), where bremsstrahlung energy can be as high as 7 GeV, production of photoneutrons in varying yields is possible from all of the above three processes. The bremsstrahlung produced along a typical 15.38-m straight path of the insertion device (ID) beamline of the APS has been measured and analyzed in previous studies. High-Z materials constituting the beamline components, such as collimators and beam stops, can produce photoneutrons upon interaction with these bremsstrahlung photons. The 1/E nature of the bremsstrahlung spectrum and the fact that the photoneutron production cross section is comparatively larger in the energy region 10 MeV γ 3 detector, as well as a very sensitive pressurized 3 He detector, is used for neutron dose measurements. The dose equivalent rates, normalized to bremsstrahlung power, beam current, and storage ring vacuum, are measured for various targets. This report details the experimental setup,

  16. SU-E-T-336: Dosimetric Properties of a New Solid Water High Equivalency Phantom for High-Energy Photon Beams

    International Nuclear Information System (INIS)

    Araki, F; Ohno, T; Onitsuka, R; Shimohigashi, Y

    2015-01-01

    Purpose: To investigate dosimetric properties in high-energy photon beams for a Solid Water High Equivalency (SWHE, SW557) phantom (Gammex) which was newly developed as water mimicking material. Methods: The mass density of SWHE and SWHE/water electron density ratio are 1.032 g/cm 3 and 1.005 according to the manufacturer information, respectively. SWHE is more water equivalent material in physical characteristics and uniformity than conventional SW457. This study calculated the relative ionization ratio of water and SWHE as a function of depth from the cavity dose in PTW30013 and Exradin A19 Farmer-type ionization chambers using Monte Caro simulations. The simulation was performed with a 10 x 10 cm 2 field at SAD of 100 cm for 4, 6, 10, 15, and 18 MV photons. The ionization ratio was also measured with the PTW30013 chamber for 6 and 15 MV photons. In addition, the overall perturbation factor of both chambers was calculated for both phantoms. Results: The relative ionization ratio curves for water and SWHE was in good agreement for all photon energies. The ionization ratio of water/SWHE for both chambers was 0.999–1.002, 0.999–1.002, 1.001–1.004, 1.004–1.007, and 1.006–1.010 at depths of over the buildup region for 4, 6, 10, 15, and 18 MV photons, respectively. The ionization ratio of water/SWHE increased up to 1% with increasing the photon energy. The measured ionization ratio of water/SWHE for 6 and 15 MV photons agreed well with calculated values. The overall perturbation factor for both chambers was 0.983–0.988 and 0.978–0.983 for water and SWHE, respectively, in a range from 4 MV to 18 MV. Conclusion: The depth scaling factor of water/SWHE was equal to unity for all photon energies. The ionization ratio of water/SWHE at a reference depth was equal to unity for 4 and 6 MV and larger up to 0.7% than unity for 18 MV

  17. Effective equivalent dose in the critical group due to release of radioactive effluents

    International Nuclear Information System (INIS)

    Santos, John W.A. dos; Varandas, Luciana R.; Souza, Denise N.; Souza, Cristiano B.F.; Lima, Sandro Leonardo N.; Mattos, Marcos Fernando M.; Moraes, Jose Adenildo T.

    2005-01-01

    To ensure that the emissions of radioactive material by liquid and gaseous pathways are below applicable limits it is necessary to evaluate the effective equivalent dose in the critical group, which is a magnitude that takes into consideration the modeling used and the terms radioactive activity source. The calculation of this dose considers each radionuclide released by the activity of Nuclear plant, liquid and gaseous by, and the sum of the values obtained is controlled so that this dose does not exceed the goals of the regulatory body, the CNEN and the goals established by the Nuclear power plant. To hit these targets various controls are used such as: controls for effluent monitors instrumentation, environmental monitoring programs, effluent release controls and dose calculation in the environment. According to the findings, it is concluded that during the period of operation of the plants, this dose is below of the required limits

  18. Use of prompt gamma emissions from polyethylene to estimate neutron ambient dose equivalent

    Energy Technology Data Exchange (ETDEWEB)

    Priyada, P.; Sarkar, P.K., E-mail: pradip.sarkar@manipal.edu

    2015-06-11

    The possibility of using measured prompt gamma emissions from polyethylene to estimate neutron ambient dose equivalent is explored theoretically. Monte Carlo simulations have been carried out using the FLUKA code to calculate the response of a high density polyethylene cylinder to emit prompt gammas from interaction of neutrons with the nuclei of hydrogen and carbon present in polyethylene. The neutron energy dependent responses of hydrogen and carbon nuclei are combined appropriately to match the energy dependent neutron fluence to ambient dose equivalent conversion coefficients. The proposed method is tested initially with simulated spectra and then validated using experimental measurements with an Am–Be neutron source. Experimental measurements and theoretical simulations have established the feasibility of estimating neutron ambient dose equivalent using measured neutron induced prompt gammas emitted from polyethylene with an overestimation of neutron dose at very low energies. - Highlights: • A new method for estimating H{sup ⁎}(10) using prompt gamma emissions from HDPE. • Linear combination of 2.2 MeV and 4.4 MeV gamma intensities approximates DCC (ICRP). • Feasibility of the method was established theoretically and experimentally. • The response of the present technique is very similar to that of the rem meters.

  19. Software for the estimation of organ equivalent and effective doses from diagnostic radiology procedures

    International Nuclear Information System (INIS)

    Osei, Ernest K; Barnett, Rob

    2009-01-01

    Diagnostic radiological imaging such as conventional radiography, fluoroscopy and computed tomography (CT) examinations will continue to provide tremendous benefits in modern healthcare. The benefit derived by the patient should far outweigh the risk associated with a properly conducted imaging examination. Nonetheless, it is very important to be able to quantify the risk associated with any radiological examination of patients, and effective dose has been considered a useful indicator of patient exposure. Quantification of the risks associated with radiological imaging is very important as such information will be helpful to physicians and their patients for comparing risks from various imaging examinations and for making informed decisions whenever there is a need for any radiological imaging. The determination of equivalent and effective doses in diagnostic radiology is of interest as a basis for estimates of risk from medical exposures. In this paper we describe a simple computer program OrgDose, which calculates the doses to 27 organs in the body and then calculates the organ equivalent and effective doses and the risk from various procedures in the radiology department including conventional radiography, fluoroscopy and computed tomography examinations. The program will be a useful tool for the medical and paramedical personnel who are involved with assessing organ and effective doses and risks from diagnostic radiology procedures.

  20. Dose of radiation enhancement, using silver nanoparticles in a human tissue equivalent gel dosimeter.

    Science.gov (United States)

    Hassan, Muhammad; Waheed, Muhammad Mohsin; Anjum, Muhammad Naeem

    2016-01-01

    To quantify the radiation dose enhancement in a human tissue-equivalent polymer gel impregnated with silver nanoparticles. The case-control study was conducted at the Bahawalpur Institute of Nuclear Medicine and Oncology, Bahawalpur, Pakistan, in January 2014. Silver nanoparticles used in this study were prepared by wet chemical method. Polymer gel was prepared by known quantity of gelatine, methacrylic acid, ascorbic acid, copper sulphate pentahydrate, hydroquinone and water. Different concentrations of silver nanoparticles were added to the gel during its cooling process. The gel was cooled in six plastic vials of 50ml each. Two vials were used as a control sample while four vials were impregnated with silver nanoparticles. After 22 hours, the vials were irradiated with gamma rays by aCobalt-60 unit. Radiation enhancement was assessed by taking magnetic resonance images of the vials. The images were analysed using Image J software. The dose enhancement factor was 24.17% and 40.49% for 5Gy and 10Gy dose respectively. The dose enhancement factor for the gel impregnated with 0.10mM silver nanoparticles was 32.88% and 51.98% for 5Gy and 10Gy dose respectively. The impregnation of a tissue-equivalent gel with silver nanoparticles resulted in dose enhancement and this effect was magnified up to a certain level with the increase in concentration of silver nanoparticles.

  1. Equivalent uniform dose concept evaluated by theoretical dose volume histograms for thoracic irradiation.

    Science.gov (United States)

    Dumas, J L; Lorchel, F; Perrot, Y; Aletti, P; Noel, A; Wolf, D; Courvoisier, P; Bosset, J F

    2007-03-01

    The goal of our study was to quantify the limits of the EUD models for use in score functions in inverse planning software, and for clinical application. We focused on oesophagus cancer irradiation. Our evaluation was based on theoretical dose volume histograms (DVH), and we analyzed them using volumetric and linear quadratic EUD models, average and maximum dose concepts, the linear quadratic model and the differential area between each DVH. We evaluated our models using theoretical and more complex DVHs for the above regions of interest. We studied three types of DVH for the target volume: the first followed the ICRU dose homogeneity recommendations; the second was built out of the first requirements and the same average dose was built in for all cases; the third was truncated by a small dose hole. We also built theoretical DVHs for the organs at risk, in order to evaluate the limits of, and the ways to use both EUD(1) and EUD/LQ models, comparing them to the traditional ways of scoring a treatment plan. For each volume of interest we built theoretical treatment plans with differences in the fractionation. We concluded that both volumetric and linear quadratic EUDs should be used. Volumetric EUD(1) takes into account neither hot-cold spot compensation nor the differences in fractionation, but it is more sensitive to the increase of the irradiated volume. With linear quadratic EUD/LQ, a volumetric analysis of fractionation variation effort can be performed.

  2. Comparison of dose calculation algorithms in slab phantoms with cortical bone equivalent heterogeneities

    International Nuclear Information System (INIS)

    Carrasco, P.; Jornet, N.; Duch, M. A.; Panettieri, V.; Weber, L.; Eudaldo, T.; Ginjaume, M.; Ribas, M.

    2007-01-01

    To evaluate the dose values predicted by several calculation algorithms in two treatment planning systems, Monte Carlo (MC) simulations and measurements by means of various detectors were performed in heterogeneous layer phantoms with water- and bone-equivalent materials. Percentage depth doses (PDDs) were measured with thermoluminescent dosimeters (TLDs), metal-oxide semiconductor field-effect transistors (MOSFETs), plane parallel and cylindrical ionization chambers, and beam profiles with films. The MC code used for the simulations was the PENELOPE code. Three different field sizes (10x10, 5x5, and 2x2 cm 2 ) were studied in two phantom configurations and a bone equivalent material. These two phantom configurations contained heterogeneities of 5 and 2 cm of bone, respectively. We analyzed the performance of four correction-based algorithms and one based on convolution superposition. The correction-based algorithms were the Batho, the Modified Batho, the Equivalent TAR implemented in the Cadplan (Varian) treatment planning system (TPS), and the Helax-TMS Pencil Beam from the Helax-TMS (Nucletron) TPS. The convolution-superposition algorithm was the Collapsed Cone implemented in the Helax-TMS. All the correction-based calculation algorithms underestimated the dose inside the bone-equivalent material for 18 MV compared to MC simulations. The maximum underestimation, in terms of root-mean-square (RMS), was about 15% for the Helax-TMS Pencil Beam (Helax-TMS PB) for a 2x2 cm 2 field inside the bone-equivalent material. In contrast, the Collapsed Cone algorithm yielded values around 3%. A more complex behavior was found for 6 MV where the Collapsed Cone performed less well, overestimating the dose inside the heterogeneity in 3%-5%. The rebuildup in the interface bone-water and the penumbra shrinking in high-density media were not predicted by any of the calculation algorithms except the Collapsed Cone, and only the MC simulations matched the experimental values within

  3. Comparison of dose calculation algorithms in slab phantoms with cortical bone equivalent heterogeneities.

    Science.gov (United States)

    Carrasco, P; Jornet, N; Duch, M A; Panettieri, V; Weber, L; Eudaldo, T; Ginjaume, M; Ribas, M

    2007-08-01

    To evaluate the dose values predicted by several calculation algorithms in two treatment planning systems, Monte Carlo (MC) simulations and measurements by means of various detectors were performed in heterogeneous layer phantoms with water- and bone-equivalent materials. Percentage depth doses (PDDs) were measured with thermoluminescent dosimeters (TLDs), metal-oxide semiconductor field-effect transistors (MOSFETs), plane parallel and cylindrical ionization chambers, and beam profiles with films. The MC code used for the simulations was the PENELOPE code. Three different field sizes (10 x 10, 5 x 5, and 2 x 2 cm2) were studied in two phantom configurations and a bone equivalent material. These two phantom configurations contained heterogeneities of 5 and 2 cm of bone, respectively. We analyzed the performance of four correction-based algorithms and one based on convolution superposition. The correction-based algorithms were the Batho, the Modified Batho, the Equivalent TAR implemented in the Cadplan (Varian) treatment planning system (TPS), and the Helax-TMS Pencil Beam from the Helax-TMS (Nucletron) TPS. The convolution-superposition algorithm was the Collapsed Cone implemented in the Helax-TMS. All the correction-based calculation algorithms underestimated the dose inside the bone-equivalent material for 18 MV compared to MC simulations. The maximum underestimation, in terms of root-mean-square (RMS), was about 15% for the Helax-TMS Pencil Beam (Helax-TMS PB) for a 2 x 2 cm2 field inside the bone-equivalent material. In contrast, the Collapsed Cone algorithm yielded values around 3%. A more complex behavior was found for 6 MV where the Collapsed Cone performed less well, overestimating the dose inside the heterogeneity in 3%-5%. The rebuildup in the interface bone-water and the penumbra shrinking in high-density media were not predicted by any of the calculation algorithms except the Collapsed Cone, and only the MC simulations matched the experimental values

  4. Deuterons at energies of 10 MeV to 1 TeV: Conversion coefficients for fluence-to-absorbed dose, equivalent dose, effective dose and gray equivalent, calculated using Monte Carlo radiation transport code MCNPX 2.7.C

    International Nuclear Information System (INIS)

    Copeland, K.; Parker, D. E.; Friedberg, W.

    2011-01-01

    Conversion coefficients were calculated for fluence-to-absorbed dose, fluence-to-equivalent dose, fluence-to-effective dose and fluence-to-gray equivalent for isotropic exposure of an adult female and an adult male to deuterons ( 2 H + ) in the energy range 10 MeV -1 TeV (0.01-1000 GeV). Coefficients were calculated using the Monte Carlo transport code MCNPX 2.7.C and BodyBuilder TM 1.3 anthropomorphic phantoms. Phantoms were modified to allow calculation of the effective dose to a Reference Person using tissues and tissue weighting factors from 1990 and 2007 recommendations of the International Commission on Radiological Protection (ICRP) and gray equivalent to selected tissues as recommended by the National Council on Radiation Protection and Measurements. Coefficients for the equivalent and effective dose incorporated a radiation weighting factor of 2. At 15 of 19 energies for which coefficients for the effective dose were calculated, coefficients based on ICRP 1990 and 2007 recommendations differed by < 3 %. The greatest difference, 47 %, occurred at 30 MeV. (authors)

  5. Universal Survival Curve and Single Fraction Equivalent Dose: Useful Tools in Understanding Potency of Ablative Radiotherapy

    International Nuclear Information System (INIS)

    Park, Clint; Papiez, Lech; Zhang Shichuan; Story, Michael; Timmerman, Robert D.

    2008-01-01

    Purpose: Overprediction of the potency and toxicity of high-dose ablative radiotherapy such as stereotactic body radiotherapy (SBRT) by the linear quadratic (LQ) model led to many clinicians' hesitating to adopt this efficacious and well-tolerated therapeutic option. The aim of this study was to offer an alternative method of analyzing the effect of SBRT by constructing a universal survival curve (USC) that provides superior approximation of the experimentally measured survival curves in the ablative, high-dose range without losing the strengths of the LQ model around the shoulder. Methods and Materials: The USC was constructed by hybridizing two classic radiobiologic models: the LQ model and the multitarget model. We have assumed that the LQ model gives a good description for conventionally fractionated radiotherapy (CFRT) for the dose to the shoulder. For ablative doses beyond the shoulder, the survival curve is better described as a straight line as predicted by the multitarget model. The USC smoothly interpolates from a parabola predicted by the LQ model to the terminal asymptote of the multitarget model in the high-dose region. From the USC, we derived two equivalence functions, the biologically effective dose and the single fraction equivalent dose for both CFRT and SBRT. Results: The validity of the USC was tested by using previously published parameters of the LQ and multitarget models for non-small-cell lung cancer cell lines. A comparison of the goodness-of-fit of the LQ and USC models was made to a high-dose survival curve of the H460 non-small-cell lung cancer cell line. Conclusion: The USC can be used to compare the dose fractionation schemes of both CFRT and SBRT. The USC provides an empirically and a clinically well-justified rationale for SBRT while preserving the strengths of the LQ model for CFRT

  6. Ambient neutron dose equivalent outside concrete vault rooms for 15 and 18 MV radiotherapy accelerators

    International Nuclear Information System (INIS)

    Martinez-ovalle, S. A.; Barquero, R.; Gomez-ros, J. M.; Lallena, A. M.

    2012-01-01

    In this work, the ambient dose equivalent, H*(10), due to neutrons outside three bunkers that house a 15- and a 18-MV Varian Clinac 2100C/D and a 15-MV Elekta Inor clinical linacs, has been calculated. The Monte Carlo code MCNPX (v. 2.5) has been used to simulate the neutron production and transport. The complete geometries including linacs and full installations have been built up according to the specifications of the manufacturers and the planes provided by the corresponding medical physical services of the hospitals where the three linacs operate. Two of these installations, those lodging the Varian linacs, have an entrance door to the bunker while the other one does not, although it has a maze with two bends. Various treatment orientations were simulated in order to establish plausible annual equivalent doses. Specifically anterior-posterior, posterior-anterior, left lateral, right lateral orientations and an additional one with the gantry rotated 30 deg. have been studied. Significant dose rates have been found only behind the walls and the door of the bunker, near the entrance and the console, with a maximum of 12 μSv h -1 . Dose rates per year have been calculated assuming a conservative workload for the three facilities. The higher dose rates in the corresponding control areas were 799 μSv y -1 , in the case of the facility which operates the 15-MV Clinac, 159 μSv y -1 , for that with the 15-MV Elekta, and 21 μSv y -1 for the facility housing the 18-MV Varian. A comparison with measurements performed in similar installations has been carried out and a reasonable agreement has been found. The results obtained indicate that the neutron contamination does not increase the doses above the legal limits and does not produce a significant enhancement of the dose equivalent calculated. When doses are below the detection limits provided by the measuring devices available today, MCNPX simulation provides an useful method to evaluate neutron dose equivalents

  7. Development of an experimental method for the determination of the dose equivalent indices for low - and medium energy X- and gamma rays

    International Nuclear Information System (INIS)

    Silva Estrada, J.J. da.

    1980-01-01

    An experimental method was developed to measure Dose Equivalent Indices for low and medium energy X-rays. A sphere was constructed to simulate the human body in accordance with ICRU Report 19 but using plexiglass instead of tissue equivalent material of density 1 g.cm -3 . Experimentally it was demonstrated that for the purpose of applied radiation protection both materials are equivalent in spite of a 18% higher density of plexiglass. CaF 2 :Mn and LiF:Mg might be utilized to determine the absorbed dose distribution within the sphere. Measurements indicate that the effective energy can be determined with an accuracy better than 15% for the energy range under consideration. Depth dose curves measured with ionization chamber compared with those of LiF:Mg showed an agreement better than 12% and in the case of CaF 2 :Mn better than 11% for all irradiation conditions used. Conversion factors in units rad R -1 measured with TLD and compared with those obtained from the literature based upon Monte Carlo calculation showed an agreement better than 23% for CaF 2 :Mn and 19% for LiF:Mg. It is concluded from these experiments that the system plexiglass sphere-TLD dosimeters might be used to measure Dose Equivalent Indices for low and medium energy photons. (Author) [pt

  8. Diamond detector in absorbed dose measurements in high-energy linear accelerator photon and electron beams.

    Science.gov (United States)

    Ravichandran, Ramamoorthy; Binukumar, John Pichy; Al Amri, Iqbal; Davis, Cheriyathmanjiyil Antony

    2016-03-08

    Diamond detectors (DD) are preferred in small field dosimetry of radiation beams because of small dose profile penumbras, better spatial resolution, and tissue-equivalent properties. We investigated a commercially available 'microdiamond' detector in realizing absorbed dose from first principles. A microdiamond detector, type TM 60019 with tandem electrometer is used to measure absorbed doses in water, nylon, and PMMA phantoms. With sensitive volume 0.004 mm3, radius 1.1mm, thickness 1 x10(-3) mm, the nominal response is 1 nC/Gy. It is assumed that the diamond detector could collect total electric charge (nC) developed during irradiation at 0 V bias. We found that dose rate effect is less than 0.7% for changing dose rate by 500 MU/min. The reproducibility in obtaining readings with diamond detector is found to be ± 0.17% (1 SD) (n = 11). The measured absorbed doses for 6 MV and 15 MV photons arrived at using mass energy absorption coefficients and stop-ping power ratios compared well with Nd, water calibrated ion chamber measured absorbed doses within 3% in water, PMMA, and nylon media. The calibration factor obtained for diamond detector confirmed response variation is due to sensitivity due to difference in manufacturing process. For electron beams, we had to apply ratio of electron densities of water to carbon. Our results qualify diamond dosimeter as a transfer standard, based on long-term stability and reproducibility. Based on micro-dimensions, we recommend these detectors for pretreatment dose verifications in small field irradiations like stereotactic treatments with image guidance.

  9. SU-E-T-409: Evaluation of Tissue Composition Effect On Dose Distribution in Radiotherapy with 6 MV Photon Beam of a Medical Linac

    Energy Technology Data Exchange (ETDEWEB)

    Ghorbani, M; Tabatabaei, Z; Noghreiyan, A Vejdani [Mashhad University of Medical Sciences, Mashhad (Iran, Islamic Republic of); Meigooni, A Soleimani [Comprehensive Cancer Center of Nevada, Las Vegas, NV (United States)

    2015-06-15

    Purpose: The aim of this study is to evaluate soft tissue composition effect on dose distribution for various soft tissues and various depths in radiotherapy with 6 MV photon beam of a medical linac. Methods: A phantom and Siemens Primus linear accelerator were simulated using MCNPX Monte Carlo code. In a homogeneous cubic phantom, six types of soft tissue and three types of tissue-equivalent materials were defined separately. The soft tissues were muscle (skeletal), adipose tissue, blood (whole), breast tissue, soft tissue (9-component) and soft tissue (4-component). The tissue-equivalent materials included: water, A-150 tissue-equivalent plastic and perspex. Photon dose relative to dose in 9-component soft tissue at various depths on the beam’s central axis was determined for the 6 MV photon beam. The relative dose was also calculated and compared for various MCNPX tallies including,F8, F6 and,F4. Results: The results of the relative photon dose in various materials relative to dose in 9-component soft tissue and using different tallies are reported in the form of tabulated data. Minor differences between dose distributions in various soft tissues and tissue-equivalent materials were observed. The results from F6 and F4 were practically the same but different with,F8 tally. Conclusion: Based on the calculations performed, the differences in dose distributions in various soft tissues and tissue-equivalent materials are minor but they could be corrected in radiotherapy calculations to upgrade the accuracy of the dosimetric calculations.

  10. Developing a single-aliquot protocol for measuring equivalent dose in biogenic carbonates

    International Nuclear Information System (INIS)

    Stirling, R.J.; Duller, G.A.T.; Roberts, H.M.

    2012-01-01

    Exploiting biogenic carbonates as thermoluminescence dosimeters requires an understanding of trap kinetics and an appropriate sequence with which to measure equivalent dose. The trap kinetics of two high temperature peaks (peaks II and III) from calcitic snail opercula have been investigated resulting in the calculation of lifetimes of 7.4 × 10 7 and 1.4 × 10 11 years for the two peaks respectively. Two measurement sequences, based upon changes in the application and measurement of a test dose, have been applied to peaks II and III, and though both methods were equally successful in dose recovery and production of a dose response curve some differences were observed. Primarily, the use of method 1 lead to dose dependant sensitivity change implying competition effects occurring during irradiation; method 2 did not experience this phenomenon. As a consequence method 2 was chosen as the most appropriate protocol for single-aliquot dating of this material. When assessing the TL behaviour of the two peaks, peak II performed poorly in dose recovery experiments recovering a dose 60–100% larger than that applied. Disproportionate growth of peak II in response to a beta dose applied prior to measurement, compared to growth following regeneration doses indicated that peak II was not suitable for use in single-aliquot protocols. However, dose recovery results for peak III were all within errors of unity of the given dose, and peak III was therefore chosen as the most appropriate peak for TL dosimetry in these single-aliquot procedures. The lifetime of charge in peak III is sufficient to date over many millions of years, and furthermore using the chosen method 2 the dose response curve has a D 0 of 3,250 ± 163 Gy allowing dating to over 3 million years.

  11. General requirements to implement the personal dose equivalent Hp(10) in Brazil

    International Nuclear Information System (INIS)

    Lopes, Amanda Gomes; Silva, Francisco Cesar Augusto da

    2017-01-01

    To update the dosimetry quantity with the international community, Brazil is changing the Individual Dose Hx to the Personal Dose Equivalent Hp(10). A bibliographical survey on the technical and administrative requirements of nine countries that use Hp(10) was carried out to obtain the most relevant ones. All of them follow IEC and ISO guidelines for technical requirements, while administrative requirements change from country to country. Based on countries experiences, this paper presents a list of important general requirements to implement Hp(10) and to prepare the Brazilian requirements according to the international scientific community. (author)

  12. General requirements to implement the personal dose equivalent Hp(10) in Brazil

    Energy Technology Data Exchange (ETDEWEB)

    Lopes, Amanda Gomes; Silva, Francisco Cesar Augusto da, E-mail: amandagl@bolsista.ird.gov.br [Instituto de Radioproteção e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil)

    2017-07-01

    To update the dosimetry quantity with the international community, Brazil is changing the Individual Dose Hx to the Personal Dose Equivalent Hp(10). A bibliographical survey on the technical and administrative requirements of nine countries that use Hp(10) was carried out to obtain the most relevant ones. All of them follow IEC and ISO guidelines for technical requirements, while administrative requirements change from country to country. Based on countries experiences, this paper presents a list of important general requirements to implement Hp(10) and to prepare the Brazilian requirements according to the international scientific community. (author)

  13. Optimization of equivalent uniform dose using the L-curve criterion

    International Nuclear Information System (INIS)

    Chvetsov, Alexei V; Dempsey, James F; Palta, Jatinder R

    2007-01-01

    Optimization of equivalent uniform dose (EUD) in inverse planning for intensity-modulated radiation therapy (IMRT) prevents variation in radiobiological effect between different radiotherapy treatment plans, which is due to variation in the pattern of dose nonuniformity. For instance, the survival fraction of clonogens would be consistent with the prescription when the optimized EUD is equal to the prescribed EUD. One of the problems in the practical implementation of this approach is that the spatial dose distribution in EUD-based inverse planning would be underdetermined because an unlimited number of nonuniform dose distributions can be computed for a prescribed value of EUD. Together with ill-posedness of the underlying integral equation, this may significantly increase the dose nonuniformity. To optimize EUD and keep dose nonuniformity within reasonable limits, we implemented into an EUD-based objective function an additional criterion which ensures the smoothness of beam intensity functions. This approach is similar to the variational regularization technique which was previously studied for the dose-based least-squares optimization. We show that the variational regularization together with the L-curve criterion for the regularization parameter can significantly reduce dose nonuniformity in EUD-based inverse planning

  14. Optimization of equivalent uniform dose using the L-curve criterion

    Energy Technology Data Exchange (ETDEWEB)

    Chvetsov, Alexei V; Dempsey, James F; Palta, Jatinder R [Department of Radiation Oncology, University of Florida, Gainesville, FL 32610-0385 (United States)

    2007-09-21

    Optimization of equivalent uniform dose (EUD) in inverse planning for intensity-modulated radiation therapy (IMRT) prevents variation in radiobiological effect between different radiotherapy treatment plans, which is due to variation in the pattern of dose nonuniformity. For instance, the survival fraction of clonogens would be consistent with the prescription when the optimized EUD is equal to the prescribed EUD. One of the problems in the practical implementation of this approach is that the spatial dose distribution in EUD-based inverse planning would be underdetermined because an unlimited number of nonuniform dose distributions can be computed for a prescribed value of EUD. Together with ill-posedness of the underlying integral equation, this may significantly increase the dose nonuniformity. To optimize EUD and keep dose nonuniformity within reasonable limits, we implemented into an EUD-based objective function an additional criterion which ensures the smoothness of beam intensity functions. This approach is similar to the variational regularization technique which was previously studied for the dose-based least-squares optimization. We show that the variational regularization together with the L-curve criterion for the regularization parameter can significantly reduce dose nonuniformity in EUD-based inverse planning.

  15. Optimization of equivalent uniform dose using the L-curve criterion.

    Science.gov (United States)

    Chvetsov, Alexei V; Dempsey, James F; Palta, Jatinder R

    2007-10-07

    Optimization of equivalent uniform dose (EUD) in inverse planning for intensity-modulated radiation therapy (IMRT) prevents variation in radiobiological effect between different radiotherapy treatment plans, which is due to variation in the pattern of dose nonuniformity. For instance, the survival fraction of clonogens would be consistent with the prescription when the optimized EUD is equal to the prescribed EUD. One of the problems in the practical implementation of this approach is that the spatial dose distribution in EUD-based inverse planning would be underdetermined because an unlimited number of nonuniform dose distributions can be computed for a prescribed value of EUD. Together with ill-posedness of the underlying integral equation, this may significantly increase the dose nonuniformity. To optimize EUD and keep dose nonuniformity within reasonable limits, we implemented into an EUD-based objective function an additional criterion which ensures the smoothness of beam intensity functions. This approach is similar to the variational regularization technique which was previously studied for the dose-based least-squares optimization. We show that the variational regularization together with the L-curve criterion for the regularization parameter can significantly reduce dose nonuniformity in EUD-based inverse planning.

  16. Biological clearance and committed dose equivalent in pulmonary region from inhaled radioaerosols for lung scanning

    Energy Technology Data Exchange (ETDEWEB)

    Soni, P.S.; Sharma, S.M.; Raghunath, B.; Somasundaram, S.

    1987-01-01

    Biological clearance half-lives (Tsub(b)) of different /sup 99/Tcsup(m)-labelled compounds from each lung have been determined, after administering the radioaerosol to normal subjects using the BARC dry aerosol generation and inhalation system. Based on these experimental clearance half-lives, the committed dose equivalent to the lungs has been computed using both the ICRP lung model and MIRD-11 values.

  17. Biological clearance and committed dose equivalent in pulmonary region from inhaled radioaerosols for lung scanning

    International Nuclear Information System (INIS)

    Soni, P.S.; Sharma, S.M.; Raghunath, B.; Somasundaram, S.

    1987-01-01

    Biological clearance half-lives (Tsub(b)) of different 99 Tcsup(m)-labelled compounds from each lung have been determined, after administering the radioaerosol to normal subjects using the BARC dry aerosol generation and inhalation system. Based on these experimental clearance half-lives, the committed dose equivalent to the lungs has been computed using both the ICRP lung model and MIRD-11 values. (author)

  18. A study of the responses of neutron dose equivalent survey meters with computer codes

    International Nuclear Information System (INIS)

    Sartori, D.E.; Beer, G.P. de

    1983-01-01

    The ANISN and DOT discrete-ordinates radiation transport codes for one and two dimensions have been proved as effective and simple techniques to study the response of dose equivalent neutron detectors. Comparisons between results of an experimental calibration of the Harwell 95/0075 survey meter and calculated results rendered satisfactory agreement, considering the different techniques and sources of error involved. Possible improvements in the methods and designs and causes of error are discussed. (author)

  19. Alternatives to dose, quality factor and dose equivalent for low level irradiation

    International Nuclear Information System (INIS)

    Sondhaus, C.A.; Bond, V.P.; Feinendegen, L.E.

    1988-01-01

    Randomly occurring energy deposition events produced by low levels of ionizing radiation interacting with tissue deliver variable amounts of energy to the sensitive target volumes within a small fraction of the cell population. A model is described in which an experimentally derived function relating event size to cell response probability operates mathematically on the microdosimetric event size distribution characterizing a given irradiation and thus determines the total fractional number of responding cells; this fraction measures the effectiveness of the given radiation. Normalizing to equal numbers of events produced by different radiations and applying this cell response or hit size effectiveness function (HSEF) should define radiation quality, or relative effectiveness, on a more nearly absolute basis than do the absorbed dose and dose evaluation, which are confounded when applied to low level irradiations. Examples using both calculation and experimental data are presented. 15 refs., 18 figs

  20. Photon and neutron dose discrimination using low pressure proportional counters with graphite and A150 walls

    International Nuclear Information System (INIS)

    Kylloenen, J.; Lindborg, L.

    2005-01-01

    Full text: The determination of both the low- and high-LET components of ambient dose equivalent in mixed fields is possible with microdosimetric methods. With the multiple-event microdosimetric variance covariance method the sum of those components are directly obtained also in pulsed beams. However, if the value of each dose component is needed a more extended analysis is required. The use of a graphite walled proportional detector in combination with a tissue-equivalent proportional counter in combination with the variance covariance method was here investigated. MCNP simulations were carried out for relevant energies to investigate the photon and neutron responses of the two detectors. The combined graphite and TEPC system, the Sievert instrument, was used for measurements at IRSN, Cadarache, in the workplace calibration fields of CANEL+, SIGMA, a Cf-252 and a moderated Cf(D 2 O,Cd) radiation field. The response of the instrument in various monoenergetic neutron fields is also known from measurements at PTB. The instrument took part in the measurement campaigns in workplace fields in the nuclear industry organized within the EVIDOS contract. The results are analyzed and the method of using a graphite detector compared with alternative methods of analysis is discussed. (author)

  1. Development of neutron dosimeter using CR-39 for measurement of ambient dose equivalent

    International Nuclear Information System (INIS)

    Maki, Daisuke; Shinozaki, Wakako; Ohguchi, Hiroyuki; Yamamoto, Takayoshi; Nakamura, Takayoshi

    2010-01-01

    A CR-39 has good advantages such as cumulative type dosimeter, small fading effect and gamma-ray insensitive. Therefore, we developed the wide energy-range environmental neutron dosimeter using eight CR-39s for area monitoring in this study. This dosimeter is made of octagonal columnar polyethylene block which height is 60 mm and bottom side is 25 mm. The dosimeter contains two types of CR-39s for fast neutron detection and slow neutron detection. Four CR-39s for fast neutron detection are used for detection of recoil protons produced by H (n, p) reactions. Four CR-39s for slow neutron detection are used with boron nitride converter to detect alpha-rays produced by 10 B (n, α) 7 Li reactions. Ambient dose equivalent is obtained by adding the number of etch-pits observed in four CR-39s for fast neutron detection to the number of etch-pits observed in four CR-39s for slow neutron detection with appropriate constants respectively. Dosimeters were irradiated with some energetic neutrons and evaluated results of ambient dose equivalent were compared with results from neutron transport calculations. Energy response of dosimeter shows good agreement with neutron fluence to ambient dose equivalent conversion coefficients. Directional dependence of dosimeter is at the same level as the rem-counter. (author)

  2. Determination of eye lenses dose equivalent in terms of Hp(3)

    International Nuclear Information System (INIS)

    Klamert, V.; Caresana, M.; Minchillo, G.; Tambussi, O.

    2002-01-01

    The Italian radioprotection legislation requires the determination of personal dose equivalent in terms of H p (10) and H p (0.07) and the determination of the eye lenses dose equivalent in terms of H p (3). Whereas the calibration of a dosemeter for the determination of H p (10) and H p (0.07) is feasible, the calibration of a dosemeter in terms of H p (3) is impossible, owing to the absence of the suitable phantom and the conversion coefficients h pk (3) from air kerma to H p (3). Using an anthropomorphic phantom for the irradiation, the aim of this work is to determine the experimental values of the conversion coefficients and to relate the result of the dosemeter worn on the forehead with the dose equivalent to the eye lenses. The study is performed in the X energy range from 30 keV to 100 keV, i.e. the one most widely used in medical practices

  3. Determination of the dose equivalents due to neutrons produced during therapeutic irradiations with a Varian CLINAC 2500; Determinacion de la dosis equivalente debida a neutrones producidos durante irradiaciones terapeuticas con un Varian CLINAC 2500

    Energy Technology Data Exchange (ETDEWEB)

    Carrillo, Ricardo E [Centro de Investigacion y de Estudios Avanzados del IPN, Unidad Merida, Yucatan (Mexico)

    1991-07-01

    This experiment it was designed to quantify that so important it is the dose equivalent deposited by the neutron flow that is generated by photonuclear reactions during therapeutic irradiations with X rays of produced high-energy for an accelerator Varian CLINAC 2500. This accelerator type is routinely used in the Department of Radiotherapy of the Hospital of the University of Wisconsin, E.U. The equivalent dose was measured in diverse towns of the room of irradiations using the activation of thin sheets of gold put in the center of plastic recipients full with water. In general, the recipients were 1 m or more than the floor and at distances still bigger than the walls. The irradiations were made using photons with the highest energy that you can select with this team - 24 MeV. The due equivalent dose to neutrons taken place here by the energy photons used they were measured and reported. (author)

  4. The simulated measurements of area and personal neutron-gamma dose equivalent in the building of HWRR

    International Nuclear Information System (INIS)

    Chen Changmao; Wen Youqin; Su Jingling; Liu Shuying; Liu Nairong

    1988-01-01

    The measuring methods and results for area and personal n-γ dose equivalent in the building of HWRR of Institute of Atomic Energy were reported. The reactor operated 4440 hours during 1985, the average themal power was 11 MW. The average area n-γ dose equivalents of the basement, experimental hall, corridors and laboratories in the building were 12.2, 11.6, 0.45 and 0.23 cSv/a, respectively. The fraction of the neutron dose equivalent in any working area was less than 21%. The average personal n-γ dose equivalent to radiation workers in the building was about 0.49 cSv/a, the γ dose equivalent was a major component. The measuring methods were compared

  5. Improvement of the equivalent sphere model for better estimates of skin or eye dose in space radiation environments

    International Nuclear Information System (INIS)

    Lin, Z.W.

    2011-01-01

    It is often useful to get a quick estimate of the dose or dose equivalent of an organ, such as blood-forming organs, the eye or the skin, in a radiation field. Sometimes an equivalent sphere is used to represent the organ for this purpose. For space radiation environments, recently it has been shown that the equivalent sphere model does not work for the eye or the skin in solar particle event environments. In this study, we improve the representation of the eye and the skin using a two-component equivalent sphere model. Motivated by the two-peak structure of the body organ shielding distribution for the eye and the skin, we use an equivalent sphere with two radius parameters, for example a partial spherical shell of a smaller thickness over a proper fraction of the full solid angle combined with a concentric partial spherical shell of a larger thickness over the rest of the full solid angle, to represent the eye or the skin. We find that using an equivalent sphere with two radius parameters instead of one drastically improves the accuracy of the estimates of dose and dose equivalent in space radiation environments. For example, in solar particle event environments the average error in the estimate of the skin dose equivalent using an equivalent sphere with two radius parameters is about 8%, while the average error of the conventional equivalent sphere model using one radius parameter is around 100%.

  6. Study of dose distribution in high energy photon beam used in radiotherapy

    International Nuclear Information System (INIS)

    Rafaravavy, R.; Raoelina Andriambololona; Bridier, A.

    2007-01-01

    The dose distribution in a medium traversed by a photon beam depends on beam energy, field size and medium nature. Percent depth dose (PDD), Dose Profile (DP) and Opening Collimator Factor (OCF) curves will be established to study this distribution. So, the PDD curves are composed by tree parts: the build-up region, the maximal dose and the quasi-equilibrium region. The maximum dose depth and the dose in depth increase with increasing photon beam energy but the dose surface decreases. The PDD increases with increasing field size.

  7. Determination of the conversion coefficient for ambient dose equivalent, H(10), from air kerma measurements

    International Nuclear Information System (INIS)

    Gonzalez J, F.; Alvarez R, J. T.

    2015-09-01

    Namely the operational magnitudes can be determined by the product of a conversion coefficient by exposure air kerma or fluence, etc. In particular in Mexico for the first time is determined the conversion coefficient (Cc) for operational magnitude Environmental Dose Equivalent H(10) by thermoluminescence dosimetry (TLD) technique. First 30 TLD-100 dosimeters are calibrated in terms of air kerma, then these dosimeters are irradiated inside a sphere ICRU type of PMMA and with the aid of theory cavity the absorbed dose in PMMA is determined at a depth of 10 mm within the sphere D PMMA (10), subsequently absorbed dose to ICRU tissue is corrected and the dose equivalent H(10) is determined. The Cc is determined as the ratio of H(10)/K a obtaining a value of 1.20 Sv Gy -1 with a u c = 3.66%, this being consistent with the published value in ISO-4037-3 of 1.20 Sv Gy -1 with a u c = 2%. (Author)

  8. Application of radio-thermoluminescence to the experimental study of the dose equivalent index

    International Nuclear Information System (INIS)

    Katz, Elvira-Beatriz

    1980-01-01

    First, the Systems of dose limitation proposed by the ICRP since 1965 are analysed. The notion of 'critical organ' is replaced by the system based on the effective dose equivalent (ICRP 77). In the same document, the 'dose equivalent index' (H I ) is suggested as a secondary limit. Here, its first definition, later modifications, advantages, and inconveniences are discussed. The second part of the study is devoted to the experimental determination of H I for X and gamma (Co 60 ) beams by radioluminescent (RTL) dosimetry. In order to take into account the different chromatic relative sensitivities of the detectors used lithium fluoride (powder and pellets), lithium borate (pellets) and calcium sulfate activated by dysprosium (powder) these latter are measured within the (10-1250) keV energy range. With the help of the preceding values and the results of the RTL measurements carried out inside the ICRU spherical phantom, the corresponding absorbed doses are deduced in order to calculate H I . (author) [fr

  9. Variability in dose-equivalent assessments for inhaled U3O8 concentrations

    International Nuclear Information System (INIS)

    Hewson, G.; Blyth, D.I.

    1985-01-01

    A potentially significant radiological hazard exists in the packaging area of uranium mills through the inhalation of airborne uranium octoxide (U 3 O 8 ). The Radiation Protection (Mining and Milling Code (1980) requires the measurement and assessment of quarterly, annual and cumulative dose equivalents for employees working in these areas. Arising through differences which exist between the abovementioned Code and ICRP 30, and assumptions of particle size and dust concentration distributions, confusion exists within Australia regarding the methods which can be used to make the required assessments. Exposure data were collected during routine monitoring at an operating mill facility and were interpreted using different methods and a range of assumptions. Results indicated the dust at this facility is characterised by an AMAD greater than 10 μ, and dust concentrations were distributed lognormally. Assumptions of a normal distribution may result in an overestimate of the dose equivalent. The importance of particle size in dose assessments using ICRP 30 techniques was highlighted. Information was masked when employee data was grouped to provide work category dose assessments. The use of ICRP 30 methods were recommended to provide uniformity throughout Australia

  10. Assessment of equivalent dose on the lens in cone beam computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, M. V. L.; Campos, P. S. F. [Federal University of Bahia, Department of Health Sciences, Salvador (Brazil); Andrade, M. E. A. [Federal University of Pernambuco, Department of Nuclear Energy, Recife (Brazil); Soares, M. R. [Federal University of Sergipe, Department of Physics, Sao Cristovao (Brazil); Batista, W. O., E-mail: marcusradiology@gmail.com [Federal Institute of Bahia, Department of Applied Sciences, 40.301-015 Salvador (Brazil)

    2014-08-15

    The Cone Beam Computed Tomography (CBCT) is presented as a useful test method for the evaluation of craniofacial structures. Among them stands the temporomandibular joint (T MJ) imaging as complementary to clinical evaluation. It must be considered that there is no reference levels established for diagnosis of this imaging modality. In this same context, recently the limit for crystalline lens was reviewed by ICRP which set new values to the equivalent dose. The aim of this study was to evaluate the kerma at the surface of the crystalline lens in T MJ CBCT and derive the equivalent dose. It was used an anthropomorphic phantom of the head and neck (manufactured by: Radiation Support Devices, model; Rs-230) containing equivalent tissue with dimensions of a typical patient. The dosimetric measurements were obtained by using seven pairs of thermoluminescent dosimetry (TLD) dosimeters (LiF: Mg, Ti) positioned on the surface of the crystalline lens, divided into two pairs (one pair for each eye) per scanner evaluated. The tomographic images were obtained in three CBCT equipment s (Kodak 9000, Gendex GXCB 500 and i-Cat). Values of equivalent dose obtained were: 5.82 mSv (Kodak 9000); 5.38 mSv (Gendex GXCB 500) and 7.98 mSv (i-Cat). These results demonstrate that for this type of procedure the doses are below the annual limit but may vary in accordance with the scanner and the exposure factors used in the image acquisition. The Gendex GXCB500 uses larger Fov and higher kV. It results in levels close to those obtained on Kodak 9000. Larger doses are associated with the i-Cat. Another factor that rises is the repetition of examinations due to positioning errors and / or patient movement, which may exceed the annual limit established by ICRP. Although the ICRP limits are not applied to medical exposures, it is advisable to consider the sensitivity of the organ. For this reason, it is concluded that doses per T MJ procedure on CBCT are below the annual limit and may vary

  11. Assessment of equivalent dose on the lens in cone beam computed tomography

    International Nuclear Information System (INIS)

    Oliveira, M. V. L.; Campos, P. S. F.; Andrade, M. E. A.; Soares, M. R.; Batista, W. O.

    2014-08-01

    The Cone Beam Computed Tomography (CBCT) is presented as a useful test method for the evaluation of craniofacial structures. Among them stands the temporomandibular joint (T MJ) imaging as complementary to clinical evaluation. It must be considered that there is no reference levels established for diagnosis of this imaging modality. In this same context, recently the limit for crystalline lens was reviewed by ICRP which set new values to the equivalent dose. The aim of this study was to evaluate the kerma at the surface of the crystalline lens in T MJ CBCT and derive the equivalent dose. It was used an anthropomorphic phantom of the head and neck (manufactured by: Radiation Support Devices, model; Rs-230) containing equivalent tissue with dimensions of a typical patient. The dosimetric measurements were obtained by using seven pairs of thermoluminescent dosimetry (TLD) dosimeters (LiF: Mg, Ti) positioned on the surface of the crystalline lens, divided into two pairs (one pair for each eye) per scanner evaluated. The tomographic images were obtained in three CBCT equipment s (Kodak 9000, Gendex GXCB 500 and i-Cat). Values of equivalent dose obtained were: 5.82 mSv (Kodak 9000); 5.38 mSv (Gendex GXCB 500) and 7.98 mSv (i-Cat). These results demonstrate that for this type of procedure the doses are below the annual limit but may vary in accordance with the scanner and the exposure factors used in the image acquisition. The Gendex GXCB500 uses larger Fov and higher kV. It results in levels close to those obtained on Kodak 9000. Larger doses are associated with the i-Cat. Another factor that rises is the repetition of examinations due to positioning errors and / or patient movement, which may exceed the annual limit established by ICRP. Although the ICRP limits are not applied to medical exposures, it is advisable to consider the sensitivity of the organ. For this reason, it is concluded that doses per T MJ procedure on CBCT are below the annual limit and may vary

  12. Contribution of the height in the ambient dose equivalent; Contribuicao da altitude no equivalente de dose ambiente

    Energy Technology Data Exchange (ETDEWEB)

    de Campos, Vicente de P.; Manzoli, Jose E.; Alipio, Osvaldo C.; Carneiro, Janete C.G.; Rodrigues Junior, Orlando, E-mail: vpcampos@ipen.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2013-07-01

    The aim of this study is to evaluate the contribution of natural sources in the ambient dose equivalent. The evaluation of the levels of environmental radiation was performed using thermoluminescent dosimeters CaSO{sub 4} of doped C, which have high sensitivity and little fading. The dosimeters were placed in five locations at different altitudes, covering the period from three to nine years depending on their location. The results were grouped according to the use and occupation of land in the vicinity of the measurement point.

  13. The use of the effective dose equivalent, Hsub(E), as a risk parameter in computed tomography

    International Nuclear Information System (INIS)

    Huda, W.; Sandison, G.A.

    1986-01-01

    This note employs the concept of the effective dose equivalent, Hsub(E) to overcome the problems of comparing the non-uniform radiation doses encountered in CT examinations with the whole-body dose-equivalent limits imposed for non-medical exposures for members of the public (5 mSv/year), or with the risks from familiar everyday activities such as smoking cigarettes or driving cars. (U.K.)

  14. Photon iso-effective dose for cancer treatment with mixed field radiation based on dose-response assessment from human and an animal model: clinical application to boron neutron capture therapy for head and neck cancer.

    Science.gov (United States)

    González, S J; Pozzi, E C C; Monti Hughes, A; Provenzano, L; Koivunoro, H; Carando, D G; Thorp, S I; Casal, M R; Bortolussi, S; Trivillin, V A; Garabalino, M A; Curotto, P; Heber, E M; Santa Cruz, G A; Kankaanranta, L; Joensuu, H; Schwint, A E

    2017-10-03

    model are compatible with the observed clinical outcome. The extension of the photon iso-effective dose model has allowed, for the first time, the determination of the photon iso-effective dose for unacceptable complications in the dose-limiting normal tissue. Finally, the formalism developed in this work to compute photon-equivalent doses can be applied to other therapies that combine mixed radiation fields, such as hadron therapy.

  15. Photon iso-effective dose for cancer treatment with mixed field radiation based on dose-response assessment from human and an animal model: clinical application to boron neutron capture therapy for head and neck cancer

    Science.gov (United States)

    González, S. J.; Pozzi, E. C. C.; Monti Hughes, A.; Provenzano, L.; Koivunoro, H.; Carando, D. G.; Thorp, S. I.; Casal, M. R.; Bortolussi, S.; Trivillin, V. A.; Garabalino, M. A.; Curotto, P.; Heber, E. M.; Santa Cruz, G. A.; Kankaanranta, L.; Joensuu, H.; Schwint, A. E.

    2017-10-01

    model are compatible with the observed clinical outcome. The extension of the photon iso-effective dose model has allowed, for the first time, the determination of the photon iso-effective dose for unacceptable complications in the dose-limiting normal tissue. Finally, the formalism developed in this work to compute photon-equivalent doses can be applied to other therapies that combine mixed radiation fields, such as hadron therapy.

  16. Reduction of outdoor and indoor ambient dose equivalent after decontamination in the Fukushima evacuation zones

    International Nuclear Information System (INIS)

    Yoshida-Ohuchi, Hiroko; Kanagami, Takashi; Naitoh, Yutaka; Kameyama, Mizuki; Hosoda, Masahiro

    2017-01-01

    One of the most urgent issues following the accident at the Fukushima Daiichi nuclear power plant (FDNPP) was the remediation of the land, in particular, for residential area contaminated by the radioactive materials discharged. In this study, the effect of decontamination on reduction of ambient dose equivalent outdoors and indoors was evaluated. The latter is essential for residents as most individuals spend a large portion of their time indoors. From December 2012 to November 2014, thirty-seven Japanese single-family detached wooden houses were investigated before and after decontamination in evacuation zones. Outdoor and indoor dose measurements (n=84 and 114, respectively) were collected based on in situ measurements using the NaI (Tl) scintillation surveymeter. The outdoor ambient dose equivalents [H"*(10)_o_u_t] ranged from 0.61 to 3.71 μSv h"-"1 and from 0.23 to 1.32 μSv h"-"1 before and after decontamination, respectively. The indoor ambient dose equivalents [H"*(10)"i"n] ranged from 0.29 to 2.53 μSv h"-"1 and from 0.16 to 1.22 μSv h"-"1 before and after decontamination, respectively. The values of reduction efficiency (RE), defined as the ratio by which the radiation dose has been reduced via decontamination, were evaluated as 0.47±0.13, 0.51±0.13, and 0.58±0.08 (average±σ) when H"*(10)_o_u_t <1.0 μSv h"-"1, 1.0 μSv h"-"1 < H"*(10)_o_u_t <2.0 μSv h"-"1, and 2.0 μSv h"-"1< H"*(10)_o_u_t, respectively, indicating the values of RE increased as H"*(10)_o_u_t increased. It was found that the values of RE were 0.53±0.12 outdoors and 0.41±0.09 indoors, respectively, indicating RE was larger outdoors than indoors. Indoor dose is essential as most individuals spend a large portion of their time indoors. The difference between outdoors and indoors should be considered carefully in order to estimate residents’ exposure dose before their returning home

  17. Reduction of outdoor and indoor ambient dose equivalent after decontamination in the Fukushima evacuation zones

    Energy Technology Data Exchange (ETDEWEB)

    Yoshida-Ohuchi, Hiroko; Kanagami, Takashi [Graduate School of Pharmaceutical Sciences, Tohoku University, Miyagi (Japan); Naitoh, Yutaka; Kameyama, Mizuki [Japan Environment Research Co., Ltd., Miyagi (Japan); Hosoda, Masahiro [Dept. of Radiological Life Sciences, Hirosaki University Graduate School of Health Sciences, Aomori (Japan)

    2017-03-15

    One of the most urgent issues following the accident at the Fukushima Daiichi nuclear power plant (FDNPP) was the remediation of the land, in particular, for residential area contaminated by the radioactive materials discharged. In this study, the effect of decontamination on reduction of ambient dose equivalent outdoors and indoors was evaluated. The latter is essential for residents as most individuals spend a large portion of their time indoors. From December 2012 to November 2014, thirty-seven Japanese single-family detached wooden houses were investigated before and after decontamination in evacuation zones. Outdoor and indoor dose measurements (n=84 and 114, respectively) were collected based on in situ measurements using the NaI (Tl) scintillation surveymeter. The outdoor ambient dose equivalents [H{sup *}(10){sub out}] ranged from 0.61 to 3.71 μSv h{sup -1} and from 0.23 to 1.32 μSv h{sup -1} before and after decontamination, respectively. The indoor ambient dose equivalents [H{sup *}(10){sup in}] ranged from 0.29 to 2.53 μSv h{sup -1} and from 0.16 to 1.22 μSv h{sup -1} before and after decontamination, respectively. The values of reduction efficiency (RE), defined as the ratio by which the radiation dose has been reduced via decontamination, were evaluated as 0.47±0.13, 0.51±0.13, and 0.58±0.08 (average±σ) when H{sup *}(10){sub out} <1.0 μSv h{sup -1}, 1.0 μSv h{sup -1} dose is essential as most individuals spend a large portion of their time indoors. The difference between outdoors and indoors should be considered carefully in order to estimate residents’ exposure dose before their returning home.

  18. Tritons at energies of 10 MeV to 1 TeV: Conversion coefficients for fluence-to-absorbed dose, equivalent dose, effective dose, and gray equivalent, calculated using Monte Carlo radiation transport code MCNPX 2.7.C

    International Nuclear Information System (INIS)

    Copeland, K.; Parker, D. E.; Friedberg, W.

    2010-01-01

    Conversion coefficients were calculated for fluence-to-absorbed dose, fluence-to-equivalent dose, fluence-to-effective dose and fluence-to-gray equivalent for isotropic exposure of an adult female and an adult male to tritons ( 3 H + ) in the energy range of 10 MeV to 1 TeV (0.01-1000 GeV). Coefficients were calculated using Monte Carlo transport code MCNPX 2.7.C and BodyBuilder TM 1.3 anthropomorphic phantoms. Phantoms were modified to allow calculation of effective dose to a Reference Person using tissues and tissue weighting factors from 1990 and 2007 recommendations of the International Commission on Radiological Protection (ICRP) and calculation of gray equivalent to selected tissues as recommended by the National Council on Radiation Protection and Measurements. At 15 of the 19 energies for which coefficients for effective dose were calculated, coefficients based on ICRP 2007 and 1990 recommendations differed by less than 3%. The greatest difference, 43%, occurred at 30 MeV. Published by Oxford Univ. Press on behalf of the US Government 2010. (authors)

  19. Helions at energies of 10 MeV to 1 TeV: Conversion coefficients for fluence-to-absorbed dose, equivalent dose, effective dose and gray equivalent, calculated using Monte Carlo radiation transport code MCNPX 2.7.C

    International Nuclear Information System (INIS)

    Copeland, K.; Parker, D. E.; Friedberg, W.

    2010-01-01

    Conversion coefficients were calculated for fluence-to-absorbed dose, fluence-to-equivalent dose, fluence-to-effective dose and fluence-to-gray equivalent, for isotropic exposure of an adult male and an adult female to helions ( 3 He 2+ ) in the energy range of 10 MeV to 1 TeV (0.01-1000 GeV). Calculations were performed using Monte Carlo transport code MCNPX 2.7.C and BodyBuilder TM 1.3 anthropomorphic phantoms modified to allow calculation of effective dose using tissues and tissue weighting factors from either the 1990 or 2007 recommendations of the International Commission on Radiological Protection (ICRP), and gray equivalent to selected tissues as recommended by the National Council on Radiation Protection and Measurements. At 15 of the 19 energies for which coefficients for effective dose were calculated, coefficients based on ICRP 2007 and 1990 recommendations differed by less than 2%. The greatest difference, 62%, occurred at 100 MeV. Published by Oxford Univ. Press on behalf of the U.S. Government 2010. (authors)

  20. A phantom for assessing the personal dose equivalent, H{sub P}(10); Um fantoma para a avaliacao do equivalente de dose pessoal, H{sub P}(10).

    Energy Technology Data Exchange (ETDEWEB)

    Santoro, C.; Filho, J.A [Universidade Federal de Pernambuco (UFPE), Recife, PE (Brazil). Departamento de Energia Nuclear; Santos, M.A.P.; Filho, L.C.G., E-mail: chsantoro@gmail.com, E-mail: masantos@cnen.gov.br, E-mail: l.filho@cnen.gov.br, E-mail: jaf@ufpe.br [Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE/CNEN-PE), Recife, PE (Brazil)

    2013-10-15

    Characteristics of a phantom designed to evaluate the personal dose equivalent, H{sub P}(10), and appropriate for photographic dosimetry are presented. It is called HP(10) phantom due to cavities constructed to insert dosimetric films at a depth of 10 mm. The H{sub P}(10) phantom is irradiated with ionizing radiation energy, E, from 45 to 1250 keV, with doses ranging from 0.2 to 50 mSv. It is positioned in the direction α = 0 °, and the radiation field focusing perpendicular to its front surface. So, are established calibration curves of dosimeters in the position conventionally true and quantities H{sub P}(10). It made a comparison between the responses obtained with the H{sub P}(10) phantom and responses obtained when using the calibration procedure recommended by ISO dosimeters. The ISO recommends getting the air kerma, Ka, for photons at test point of the radiation field by an ionization chamber. And through conversion coefficients, h{sub pK} (10; E, α), becomes the air kerma for H{sub P}(10). The ISO 4037-3 recommendation has been studied by researchers to ensure that the low energy spectral differences occur in radiation fields which are generated by various X-ray equipment, and induce changes in the percentages of conversion coefficients on the order of 10% to 90% . On the basis of the recommendations ISO, this article develops phantom able to assess the dose to the influence of scattering and absorption of radiation, its implications with respect to dosimetry, providing improvement in the assessment of doses. (author)

  1. Neutron spectrum and dose-equivalent in shuttle flights during solar maximum

    Energy Technology Data Exchange (ETDEWEB)

    Keith, J E; Badhwar, G D; Lindstrom, D J [National Aeronautics and Space Administration, Houston, TX (United States). Lyndon B. Johnson Space Center

    1992-01-01

    This paper presents unambiguous measurements of the spectrum of neutrons found in spacecraft during spaceflight. The neutron spectrum was measured from thermal energies to about 10 MeV using a completely passive system of metal foils as neutron detectors. These foils were exposed to the neutron flux bare, covered by thermal neutron absorbers (Gd) and inside moderators (Bonner spheres). This set of detectors was flown on three U.S. Space Shuttle flights, STS-28, STS-36 and STS-31, during the solar maximum. We show that the measurements of the radioactivity of these foils lead to a differential neutron energy spectrum in all three flights that can be represented by a power law, J(E){approx equal}E{sup -0.765} neutrons cm{sup -2} day {sup -1} MeV{sup -1}. We also show that the measurements are even better represented by a linear combination of the terrestrial neutron albedo and a spectrum of neutrons locally produced in a aluminium by protons, computed by a previous author. We use both approximations to the neutron spectrum to produce a worst case and most probable case for the neutron spectra and the resulting dose-equivalents, computed using ICRP-51 neutron fluence-dose conversion tables. We compare these to the skin dose-equivalents due to charged particles during the same flights. (author).

  2. Test of tissue-equivalent scintillation detector for dose measurement of megavoltage beams

    International Nuclear Information System (INIS)

    Geso, M.; Ackerly, T.; Clift, M.A.

    2000-01-01

    Full text: The measurement of depth doses and profiles for a stereotactic radiotherapy beam presents special problems associated with the small beam size compared to the dosimeter's active detection area. In this work a locally fabricated organic plastic scintillator detector has been used to measure the depth dose and profile of a stereotactic radiotherapy beam. The 6MV beam is 1.25 cm diameter at isocentre, typical of small field stereotactic radiosurgery. The detector is a water/tissue equivalent plastic scintillator that is accompanied by Cerenkov subtraction detector. In this particular application, a negligible amount of Cerenkov light was detected. A photodiode and an electronic circuit is used instead of a photomultiplier for signal amplification. Comparison with data using a diode detector and a small size ionization chamber, indicate that the organic plastic scintillator detector is a valid detector for stereotactic radiosurgery dosimetry. The tissue equivalence of the organic scintillator also holds the promise of accurate dosimetry in the build up region. Depth doses measured using our plastic scintillator agree to within about 1% with those obtained using commercially available silicon diodes. Beam profiles obtained using plastic scintillator presents correct field width to within 0.35 mm, however some artifacts are visible in the profiles. These artifacts are about 5% discrepancy which has been shown not to be a significant factor in stereotactic radiotherapy dosimetry. Copyright (2000) Australasian College of Physical Scientists and Engineers in Medicine

  3. Comparison between the calculated and measured dose distributions for four beams of 6 MeV linac in a human-equivalent phantom

    Directory of Open Access Journals (Sweden)

    Reda Sonia M.

    2006-01-01

    Full Text Available Radiation dose distributions in various parts of the body are of importance in radiotherapy. Also, the percent depth dose at different body depths is an important parameter in radiation therapy applications. Monte Carlo simulation techniques are the most accurate methods for such purposes. Monte Carlo computer calculations of photon spectra and the dose ratios at surfaces and in some internal organs of a human equivalent phantom were performed. In the present paper, dose distributions in different organs during bladder radiotherapy by 6 MeV X-rays were measured using thermoluminescence dosimetry placed at different points in the human-phantom. The phantom was irradiated in exactly the same manner as in actual bladder radiotherapy. Four treatment fields were considered to maximize the dose at the center of the target and minimize it at non-target healthy organs. All experimental setup information was fed to the MCNP-4b code to calculate dose distributions at selected points inside the proposed phantom. Percent depth dose distribution was performed. Also, the absorbed dose as ratios relative to the original beam in the surrounding organs was calculated by MCNP-4b and measured by thermoluminescence dosimetry. Both measured and calculated data were compared. Results indicate good agreement between calculated and measured data inside the phantom. Comparison between MCNP-4b calculations and measurements of depth dose distribution indicated good agreement between both.

  4. A new online detector for estimation of peripheral neutron equivalent dose in organ

    Energy Technology Data Exchange (ETDEWEB)

    Irazola, L., E-mail: leticia@us.es; Sanchez-Doblado, F. [Departamento de Fisiología Médica y Biofísica, Universidad de Sevilla, Sevilla 41009, Spain and Servicio de Radiofísica, Hospital Universitario Virgen Macarena, Sevilla 41007 (Spain); Lorenzoli, M.; Pola, A. [Departimento di Ingegneria Nuclear, Politecnico di Milano, Milano 20133 (Italy); Bedogni, R. [Laboratori Nazionali di Frascati, Istituto Nazionale di Fisica Nucleare (INFN), Frascati Roma 00044 (Italy); Terrón, J. A. [Servicio de Radiofísica, Hospital Universitario Virgen Macarena, Sevilla 41007 (Spain); Sanchez-Nieto, B. [Instituto de Física, Pontificia Universidad Católica de Chile, Santiago 4880 (Chile); Expósito, M. R. [Departamento de Física, Universitat Autònoma de Barcelona, Bellaterra 08193 (Spain); Lagares, J. I.; Sansaloni, F. [Centro de Investigaciones Energéticas y Medioambientales y Tecnológicas (CIEMAT), Madrid 28040 (Spain)

    2014-11-01

    Purpose: Peripheral dose in radiotherapy treatments represents a potential source of secondary neoplasic processes. As in the last few years, there has been a fast-growing concern on neutron collateral effects, this work focuses on this component. A previous established methodology to estimate peripheral neutron equivalent doses relied on passive (TLD, CR39) neutron detectors exposed in-phantom, in parallel to an active [static random access memory (SRAMnd)] thermal neutron detector exposed ex-phantom. A newly miniaturized, quick, and reliable active thermal neutron detector (TNRD, Thermal Neutron Rate Detector) was validated for both procedures. This first miniaturized active system eliminates the long postprocessing, required for passive detectors, giving thermal neutron fluences in real time. Methods: To validate TNRD for the established methodology, intrinsic characteristics, characterization of 4 facilities [to correlate monitor value (MU) with risk], and a cohort of 200 real patients (for second cancer risk estimates) were evaluated and compared with the well-established SRAMnd device. Finally, TNRD was compared to TLD pairs for 3 generic radiotherapy treatments through 16 strategic points inside an anthropomorphic phantom. Results: The performed tests indicate similar linear dependence with dose for both detectors, TNRD and SRAMnd, while a slightly better reproducibility has been obtained for TNRD (1.7% vs 2.2%). Risk estimates when delivering 1000 MU are in good agreement between both detectors (mean deviation of TNRD measurements with respect to the ones of SRAMnd is 0.07 cases per 1000, with differences always smaller than 0.08 cases per 1000). As far as the in-phantom measurements are concerned, a mean deviation smaller than 1.7% was obtained. Conclusions: The results obtained indicate that direct evaluation of equivalent dose estimation in organs, both in phantom and patients, is perfectly feasible with this new detector. This will open the door to an

  5. Assessment of doses caused by electrons in thin layers of tissue-equivalent materials, using MCNP.

    Science.gov (United States)

    Heide, Bernd

    2013-10-01

    Absorbed doses caused by electron irradiation were calculated with Monte Carlo N-Particle transport code (MCNP) for thin layers of tissue-equivalent materials. The layers were so thin that the calculation of energy deposition was on the border of the scope of MCNP. Therefore, in this article application of three different methods of calculation of energy deposition is discussed. This was done by means of two scenarios: in the first one, electrons were emitted from the centre of a sphere of water and also recorded in that sphere; and in the second, an irradiation with the PTB Secondary Standard BSS2 was modelled, where electrons were emitted from an (90)Sr/(90)Y area source and recorded inside a cuboid phantom made of tissue-equivalent material. The speed and accuracy of the different methods were of interest. While a significant difference in accuracy was visible for one method in the first scenario, the difference in accuracy of the three methods was insignificant for the second one. Considerable differences in speed were found for both scenarios. In order to demonstrate the need for calculating the dose in thin small zones, a third scenario was constructed and simulated as well. The third scenario was nearly equal to the second one, but a pike of lead was assumed to be inside the phantom in addition. A dose enhancement (caused by the pike of lead) of ∼113 % was recorded for a thin hollow cylinder at a depth of 0.007 cm, which the basal-skin layer is referred to in particular. Dose enhancements between 68 and 88 % were found for a slab with a radius of 0.09 cm for all depths. All dose enhancements were hardly noticeable for a slab with a cross-sectional area of 1 cm(2), which is usually applied to operational radiation protection.

  6. Superiority of Equivalent Uniform Dose (EUD)-Based Optimization for Breast and Chest Wall

    International Nuclear Information System (INIS)

    Mihailidis, Dimitris N.; Plants, Brian; Farinash, Lloyd; Harmon, Michael; Whaley, Lewis; Raja, Prem; Tomara, Pelagia

    2010-01-01

    We investigate whether IMRT optimization based on generalized equivalent uniform dose (gEUD) objectives for organs at risk (OAR) results in superior dosimetric outcomes when compared with multiple dose-volume (DV)-based objectives plans for patients with intact breast and postmastectomy chest wall (CW) cancer. Four separate IMRT plans were prepared for each of the breast and CW cases (10 patients). The first three plans used our standard in-house, physician-selected, DV objectives (phys-plan); gEUD-based objectives for the OARs (gEUD-plan); and multiple, 'very stringent,' DV objectives for each OAR and PTV (DV-plan), respectively. The fourth plan was only beam-fluence optimized (FO-plan), without segmentation, which used the same objectives as in the DV-plan. The latter plan was to be used as an 'optimum' benchmark without the effects of the segmentation for deliverability. Dosimetric quantities, such as V 20Gy for the ipsilateral lung and mean dose (D mean ) for heart, contralateral breast, and contralateral lung were used to evaluate the results. For all patients in this study, we have seen that the gEUD-based plans allow greater sparing of the OARs while maintaining equivalent target coverage. The average ipsilateral lung V 20Gy reduced from 22 ± 4.4% for the FO-plan to 18 ± 3% for the gEUD-plan. All other dosimetric quantities shifted towards lower doses for the gEUD-plan. gEUD-based optimization can be used to search for plans of different DVHs with the same gEUDs. The use of gEUD allows selective optimization and reduction of the dose for each OAR and results in a truly individualized treatment plan.

  7. A new online detector for estimation of peripheral neutron equivalent dose in organ

    International Nuclear Information System (INIS)

    Irazola, L.; Sanchez-Doblado, F.; Lorenzoli, M.; Pola, A.; Bedogni, R.; Terrón, J. A.; Sanchez-Nieto, B.; Expósito, M. R.; Lagares, J. I.; Sansaloni, F.

    2014-01-01

    Purpose: Peripheral dose in radiotherapy treatments represents a potential source of secondary neoplasic processes. As in the last few years, there has been a fast-growing concern on neutron collateral effects, this work focuses on this component. A previous established methodology to estimate peripheral neutron equivalent doses relied on passive (TLD, CR39) neutron detectors exposed in-phantom, in parallel to an active [static random access memory (SRAMnd)] thermal neutron detector exposed ex-phantom. A newly miniaturized, quick, and reliable active thermal neutron detector (TNRD, Thermal Neutron Rate Detector) was validated for both procedures. This first miniaturized active system eliminates the long postprocessing, required for passive detectors, giving thermal neutron fluences in real time. Methods: To validate TNRD for the established methodology, intrinsic characteristics, characterization of 4 facilities [to correlate monitor value (MU) with risk], and a cohort of 200 real patients (for second cancer risk estimates) were evaluated and compared with the well-established SRAMnd device. Finally, TNRD was compared to TLD pairs for 3 generic radiotherapy treatments through 16 strategic points inside an anthropomorphic phantom. Results: The performed tests indicate similar linear dependence with dose for both detectors, TNRD and SRAMnd, while a slightly better reproducibility has been obtained for TNRD (1.7% vs 2.2%). Risk estimates when delivering 1000 MU are in good agreement between both detectors (mean deviation of TNRD measurements with respect to the ones of SRAMnd is 0.07 cases per 1000, with differences always smaller than 0.08 cases per 1000). As far as the in-phantom measurements are concerned, a mean deviation smaller than 1.7% was obtained. Conclusions: The results obtained indicate that direct evaluation of equivalent dose estimation in organs, both in phantom and patients, is perfectly feasible with this new detector. This will open the door to an

  8. Measurements of thorium-B (212Pb) in the outdoor environment and evaluation of equivalent dose

    International Nuclear Information System (INIS)

    Mohammed, A.; El-Hussein, A.; Ali, A.E.

    2000-01-01

    The activity size distribution of unattached as well as attached 212 Pb to aerosol particles was measured in the open air of El-Minia City, Egypt. The samples were collected using a wire screen diffusion battery technique and a low pressure cascade impactor. The mean activity median thermodynamic diameter (AMTD) of unattached 212 Pb was determined to be 1.4 nm with a relative mean geometric standard deviation (σ g ) of 1.55. A mean unattached fraction (f p ) of 0.03±0.007 was obtained at a mean aerosol particle concentration of 32x10 3 cm -3 . Sometimes the f p values were less than the detection limit of 0.008. The mean concentration of activity of 212 Pb was found to be 9.6±1.1 mBq m -3 . The mean activity median aerodynamic diameter (AMAD) of the accumulation mode of attached 212 Pb was determined to be 360 nm with a mean (σ g ) of 2.7. The mean value of specific air activity concentration of 212 Pb associated with that mode was determined to be 303±12 mBq m -3 . With a dosimetric model calculation (International Commission on Radiological Protection. Human respiratory tract model for radiological protection. Oxford: Pergamon Press, ICRP Publication 66, 1994) the total deposition fractions as well as total equivalent and effective dose have been evaluated considering the obtained parameters of the activity size distributions. At a total deposition fraction of about 97% for unattached activities the total equivalent and effective doses to the lung were determined to be about 0.18 and 0.02 μSv, respectively, while total equivalent and effective doses of about 0.45 and 0.05 μ Sv, respectively, were determined at a total deposition fraction of about 23% for the attached activities

  9. Evaluation of the collective effective dose equivalent of atmospheric discharges in the Rhone valley

    International Nuclear Information System (INIS)

    Le Grand, J.; Despres, A.; Robeau, D.; Bouville, A.

    1982-01-01

    Two models were used to evaluate the integrated atmospheric concentrations: - TALD, a long-range atmospheric transport model which takes account of the actual trajectories of the discharged matter; - a Gaussian plume model which assumes straight-line trajectories defined by the wind at the point of emission. In order to bring out the differences due to the use of the two models the dose equivalents are presented as a function of wind direction at the point of emission and the contributions of the near zone and the further zone are compared. (author)

  10. Feasibility of optimizing the dose distribution in lung tumors using fluorine-18-fluorodeoxyglucose positron emission tomography and single photon emission computed tomography guided dose prescriptions

    International Nuclear Information System (INIS)

    Das, S.K.; Miften, M.M.; Zhou, S.; Bell, M.; Munley, M.T.; Whiddon, C.S.; Craciunescu, O.; Baydush, A.H.; Wong, T.; Rosenman, J.G.; Dewhirst, M.W.; Marks, L.B.

    2004-01-01

    The information provided by functional images may be used to guide radiotherapy planning by identifying regions that require higher radiation dose. In this work we investigate the dosimetric feasibility of delivering dose to lung tumors in proportion to the fluorine-18-fluorodeoxyglucose activity distribution from positron emission tomography (FDG-PET). The rationale for delivering dose in proportion to the tumor FDG-PET activity distribution is based on studies showing that FDG uptake is correlated to tumor cell proliferation rate, which is shown to imply that this dose delivery strategy is theoretically capable of providing the same duration of local control at all voxels in tumor. Target dose delivery was constrained by single photon emission computed tomography (SPECT) maps of normal lung perfusion, which restricted irradiation of highly perfused lung and imposed dose-function constraints. Dose-volume constraints were imposed on all other critical structures. All dose-volume/function constraints were considered to be soft, i.e., critical structure doses corresponding to volume/function constraint levels were minimized while satisfying the target prescription, thus permitting critical structure doses to minimally exceed dose constraint levels. An intensity modulation optimization methodology was developed to deliver this radiation, and applied to two lung cancer patients. Dosimetric feasibility was assessed by comparing spatially normalized dose-volume histograms from the nonuniform dose prescription (FDG-PET proportional) to those from a uniform dose prescription with equivalent tumor integral dose. In both patients, the optimization was capable of delivering the nonuniform target prescription with the same ease as the uniform target prescription, despite SPECT restrictions that effectively diverted dose from high to low perfused normal lung. In one patient, both prescriptions incurred similar critical structure dosages, below dose-volume/function limits

  11. The study on pre-heat conditions in the equivalent-dose estimation of holocene loess using the single-aliquot regenerative-dose (SAR) protocol

    International Nuclear Information System (INIS)

    Jia Yaofeng; Huang Chunchang; Pang Jiangli; Lu Xinwei; Zhang Xu

    2008-01-01

    The thermal treatment in the equivalent-dose estimation often is carried in the OSL dating, and pre-heat is a main thermal treatment. Due to which will originate the problems of thermal transfer and thermal activation, the thermal treatment and the setup of their conditions are key problems influencing the accuracy of OSL dating. The paper combined the temperature of pre-heat and cut-heat used in the routine measurement of IRSL and Post-IR OSL, and then estimated the equivalent-dose of several loess samples. The estimated result presents that the equivalent-dose depends on the heat temperature, especially depends on the cut-heat temperature, which is to say that the equivalent-dose increases with the cut-heat temperature; a plateau of equivalent-dose appears when using the 200-240 degree C cut-heat in the range of 200-300 degree C pre-heat, and the equivalent-doses estimated by IRSL and Post-IR OSL respectively are close to each other, which resulted from the similar sensitivity change direction of optical stimulated signals and its smaller change range in the measurement cycles using the combined temperature of pre- heat and cut-heat, and the incomplete calibration of sensitivity change of optical stimulated signals in the whole measurement cycles caused the variation of estimated equivalent-dose corresponding to the cut-heat temperature. (authors)

  12. Comment on ''Modified photon equation of motion as a test for the principle of equivalence''

    International Nuclear Information System (INIS)

    Nityananda, R.

    1992-01-01

    In a recent paper, a modification of the geodesic equation was proposed for spinning photons containing a spin-curvature coupling term. The difference in arrival times of opposite circular polarizations starting simultaneously from a source was computed, obtaining a result linear in the coupling parameter. It is pointed out here that this linear term violates causality and, more generally, Fermat's principle, implying calculational errors. Even if these are corrected, there is a violation of covariance in the way the photon spin was introduced. Rectifying this makes the effect computed vanish entirely

  13. PUDEQ: a computer code for calculating dose equivalent from internal deposition of plutonium at Hanford

    International Nuclear Information System (INIS)

    Houston, J.R.; Heid, K.R.

    1975-10-01

    Presented here are the procedures and mathematical models used in developing PUDEQ, a computer program for computing the dose equivalent to body organs from intake of Pu. The program was designed specifically to use the data recorded on the Hanford Internal Exposure (HIE) System magnetic tape as input. Insofar as was possible, the recommendations of the Advisory Committee on Dose from Plutonium and other Transuranics was followed. Some deviations were made where errors, omissions, or inconsistencies were found, after consultation with members of the Committee. In the current version of the program only Pu and its immediate important daughters are considered. The program could, however, be expanded to include other transuranic nuclides. At present, only a few depositions of transuranic nuclides other than plutonium are recorded out of about 450 individuals involved in a total of over 700 plutonium intakes

  14. Bioequivalence of a fixed-dose repaglinide/metformin combination tablet and equivalent doses of repaglinide and metformin tablets
.

    Science.gov (United States)

    Cho, Hea-Young; Ngo, Lien; Kim, Sang-Ki; Choi, Yoonho; Lee, Yong-Bok

    2018-06-01

    This study was conducted to determine whether a fixed-dose combination (FDC) tablet of repaglinide/metformin (2/500 mg) is equivalent to coadministration of equivalent doses of individual (EDI) tablets of repaglinide (2 mg) and metformin (500 mg) in healthy Korean male subjects. This study was conducted as an open-label, randomized, single-dose, two-period, two-sequence crossover design in 50 healthy Korean male subjects who received an FDC tablet or EDI tablets. Plasma concentrations of repaglinide and metformin were determined for up to 24 hours using a validated UPLC-MS/MS method. Bioequivalence was assessed according to current guidelines issued by the U.S. Food and Drug Administration (FDA) and Korean legislation. Tolerability was also evaluated throughout the study via subject interview, vital signs, and blood sampling. Point estimates (90% CIs) for AUC0-t, AUC0-∞, and Cmax based on EDI tablets were 110.07 (102.25 - 118.49), 109.90 (101.70 - 118.39), and 112.60 (101.49 - 124.85), respectively, for repaglinide. They were 95.18 (89.62 - 101.05), 95.00 (89.74 - 100.65), and 98.44 (92.72 - 104.50), respectively, for metformin. These results satisfied the bioequivalence criteria of 80.00 - 125.00% proposed by the FDA and Korean legislation. Results of pharmacokinetic analysis suggested that repaglinide and metformin in FDC tablets were bioequivalent to EDI tablets of repaglinide (2 mg) and metformin (500 mg) in healthy Korean male subjects. Both formulations appeared to be well tolerated.
.

  15. The intercomparison of the dose distributions between conformation techniques with pions and photons

    International Nuclear Information System (INIS)

    Karasawa, K.; Nakagawa, K.; Akanuma, A.

    1990-01-01

    To compare conformation radiation treatment with pions vs photons, dose volume histograms (DVH) to the critical organs, including the spinal cord, kidney, and intestine, were examined in a patient with retroperitoneal soft tissue sarcoma. For photon conformation treatment, the following techniques were used: 360 degree rotation conformation technique (photon conformation), 4 fixed field technique (photon 4-field), and 2-axis conformation technique (photon 2-axial conformation). According to the DVH reduction method, complication probability was estimated. The concave portion of the target was conformed by pion conformation treatment, but not by photon conformation treatment. Pion conformation for the intestine showed the best DVH, whereas photon 4-field technique showed the worst DVH. For the kidney, pion conformation showed better DVH as compared with any other photon conformation treatment technique. In the spinal cord, photon 2-axial conformation was far superior, followed by pion conformation and then photon conformation and 4-field technique. A 2-axial technique showed a bigger inhomogeneity inside the target volume which is critical in curative treatment. TD 50 was 72 Gy for pion conformation, 53 Gy for photon conformation, 51 Gy for photon 4-field, and 68 Gy for photon 2-axial conformation. Complication probabilities for these conformation techniques at 60 Gy were 3%, 85%, 97%, and 9%. In view of tumor control probabilities, pion seems to have the biggest therapeutic ratio among these techniques. (N.K.)

  16. Efficiency improvement in proton dose calculations with an equivalent restricted stopping power formalism

    Science.gov (United States)

    Maneval, Daniel; Bouchard, Hugo; Ozell, Benoît; Després, Philippe

    2018-01-01

    The equivalent restricted stopping power formalism is introduced for proton mean energy loss calculations under the continuous slowing down approximation. The objective is the acceleration of Monte Carlo dose calculations by allowing larger steps while preserving accuracy. The fractional energy loss per step length ɛ was obtained with a secant method and a Gauss-Kronrod quadrature estimation of the integral equation relating the mean energy loss to the step length. The midpoint rule of the Newton-Cotes formulae was then used to solve this equation, allowing the creation of a lookup table linking ɛ to the equivalent restricted stopping power L eq, used here as a key physical quantity. The mean energy loss for any step length was simply defined as the product of the step length with L eq. Proton inelastic collisions with electrons were added to GPUMCD, a GPU-based Monte Carlo dose calculation code. The proton continuous slowing-down was modelled with the L eq formalism. GPUMCD was compared to Geant4 in a validation study where ionization processes alone were activated and a voxelized geometry was used. The energy straggling was first switched off to validate the L eq formalism alone. Dose differences between Geant4 and GPUMCD were smaller than 0.31% for the L eq formalism. The mean error and the standard deviation were below 0.035% and 0.038% respectively. 99.4 to 100% of GPUMCD dose points were consistent with a 0.3% dose tolerance. GPUMCD 80% falloff positions (R80 ) matched Geant’s R80 within 1 μm. With the energy straggling, dose differences were below 2.7% in the Bragg peak falloff and smaller than 0.83% elsewhere. The R80 positions matched within 100 μm. The overall computation times to transport one million protons with GPUMCD were 31-173 ms. Under similar conditions, Geant4 computation times were 1.4-20 h. The L eq formalism led to an intrinsic efficiency gain factor ranging between 30-630, increasing with the prescribed accuracy of simulations. The

  17. Evaluation of the directional dose equivalent H,(0.07) for ring dosemeters

    International Nuclear Information System (INIS)

    Alvarez R, J.T.; Tovar M, V.M.

    2006-01-01

    The personnel dosimetry laboratory (LDP) of the Metrology department received an user's of radiation beta application that incidentally had irradiated 14 couples of ring dosemeters for extremities of the type TLD-100 given by the LDP. This sample of 14 couples of rings tentatively it was irradiated in the months of July-August of the year 2004, and he requested in an expedite way the evaluation of the received dose equivalent. The LSCD builds two calibration curves in terms of the directional dose equivalent H'(0.07) using two sources patterns of 90 Sr- 90 Y for beta radiation: one of 74 MBq and another of 1850 MBq with traceability to the PTB. The first curve in the interval of 0 to 5 mSv, the second in the range of 5 to 50 mSv, taking into account effects by positioned of the rings in the phantom. Both calibration curves were validated by adjustment lack, symmetry of residuals and normality of the same ones. It is evaluated and analyzing the H'(0.007) for these 14 couples of rings using the Tukey test of media of a single road. It was found that the H , its could be classified in 4 groups, and that the probability that its has irradiated in a random way it was smaller to the level at α = 0.05. (Author)

  18. Individual monitoring of external exposure in terms of personal dose equivalent, Hp(d)

    International Nuclear Information System (INIS)

    Fantuzzi, E.

    2001-01-01

    The institute for Radiation Protection of ENEA - Bologna has organised a one day-workshop on the subject: Individual monitoring of external exposure in terms of personal dose equivalent, H p (d). The aim of the workshop was the discussion of the new implications and modifications to be expected in the routine individual monitoring of external radiation, due to the issue of the Decree 241/00 (G.U. 31/8/2000) in charge since 01/01/2001. The decree set up in Italian law the standards contained in the European Directive EURATOM 96/29-Basic Standards for the Protection of Health of Workers and the General Public against Dangers arising from Ionizing Radiation. Among others, the definition of the operational quantities for external radiation for personal and environmental monitoring, H p (d) e H * (d) respectively as defined by ICRU (International Commission for Radiation Units and Measurements), requires to update the methods of measurements and calibration of the personal dosemeters and environmental monitors. This report collects the papers presented at the workshop dealing with the Personal Dose Equivalent, H p (d), the conversion coefficients, H p (d)/K a e H p (d)/ , obtained through Monte Carlo calculations published by ICRU and ICRP (International Commission for Radiation Protection), the new calibration procedures and the practical implication in the routine of individual monitoring in terms of H p (d). Eventually, in the last chapter, the answers to Frequently Asked Questions (FAQ) are briefly reported [it

  19. Angular dependence of dose equivalent response of an albedo neutron dosimeter

    International Nuclear Information System (INIS)

    Torres, B.A.; Boswell, E.; Schwartz, R.B.

    1994-01-01

    The ANSI provides procedures for testing the performance of dosimetry services. Although neutron dose equivalent angular response studies are not now mandated, future standards may well require that such studies be performed. Current studies with an albedo dosimeter will yield information regarding the angular dependence of dose equivalent response for this type of personnel dosimeter. Preliminary data for bare 252 Cf fluences show a marked decrease in dosimeter reading with increasing angle. The response decreased by an approximate factor of four. For the horizontal orientation, the same response was noted from both positive and negative angles. However, for the vertical orientation, the response was unexplainably assymetric. We are also examining the response of the personnel badge in moderated 252 Cf fluences. Responses from the moderated and unmoderated 252 Cf fields and theoretical calculations of the neutron angular response will be compared. This information will assist in building a data base for future comparisons of neutron angular responses with other neutron albedo dosimeters and phantoms

  20. The monetary value of the collective dose equivalent unit (person-rem)

    International Nuclear Information System (INIS)

    Rodgers, Reginald C.

    1978-01-01

    In the design and operation of nuclear power reactor facilities, it is recommended that radiation exposures to the workers and the general public be kept as 'low as reasonably achievable' (ALARA). In the process of implementing this principle cost-benefit evaluations are part of the decision making process. For this reason a monetary value has to be assigned to the collective dose equivalent unit (person-rem). The various factors such as medical health care, societal penalty and manpower replacement/saving are essential ingredients to determine a monetary value for the person-rem. These factors and their dependence on the level of risk (or exposure level) are evaluated. Monetary values of well under $100 are determined for the public dose equivalent unit. The occupational worker person-rem value is determined to be in the range of $500 to about $5000 depending on the exposure level and the type of worker and his affiliation, i.e., temporary or permanent. A discussion of the variability and the range of the monetary values will be presented. (author)

  1. Comparison between measured and design dose rate equivalents on board of Nuclear Ship Mutsu

    International Nuclear Information System (INIS)

    Yamaji, Akio; Sakamoto, Yukio

    1993-01-01

    The power-up test of the Nuclear Ship Mutsu was restarted in March 1990 and completed successfully in February 1991. The experimental voyages were carried out for about one year and all experiments were completed in February 1992. A comparison between the measured and design dose rate equivalents on board is described with showing a modified method in the shielding design. The measured values were obtained extensively in the cavity between the primary and secondary shields, in the double bottom, outside the secondary shield, and on the surface of the main coolant loop. The shielding design calculations were made with the most conservative geometries and material compositions within the allowed tolerance. In addition, a conservative model was adopted in case of performing the approximation due to the geometrical restriction of calculation code. The computational accuracies were evaluated based on various experimental analyses. The evaluated value was used as the design value. The shield structures were determined with a judgement that the real value does not exceed the design value. The adequacy of the judgement was confirmed by measurements on board. The measured dose rate equivalents in all positions on board satisfied the design criteria. (author)

  2. Equivalent dose measurements on board an Armenian Airline flight and Concord (9-17 km)

    International Nuclear Information System (INIS)

    Akopova, A.B.; Melkonyan, A.A.; Tatikyan, S.Sh.; Capdevielle, J-N.

    2002-01-01

    The results of investigations of the neutron component (E=1-10 MeV) of cosmic radiation on board the 'Armenian Airlines' aircrafts using nuclear photoemulsion are presented. The emulsions were exposed on the flights from Yerevan to Moscow, St.-Petersburg, Beirut, Athens, Frankfurt, Amsterdam, Paris and Sofia, and on Concord supersonic flights from Paris to New York. The dependence of the neutron fluxes, and on absorbed and equivalent doses on the flight parameters were investigated. On the flights of the supersonic Concord, with an altitude of 17 km, the neutron fluxes were essentially higher in comparison to those measured on Armenian airliners. It is interesting to note, that the neutron flux and equivalent dose rate decrease with altitude up to 470 km in space, for example, on board the STS-57. The shape of the differential energy spectrum for fast neutrons is the same on all Armenian airlines flights, but significantly different at 17 km altitude, where the flux in the energy region above 3 MeV is increasing

  3. Equivalent dose measurements on board an Armenian Airline flight and Concord (9-17 km)

    Energy Technology Data Exchange (ETDEWEB)

    Akopova, A.B. E-mail: akopova@lx2.yerphi.am; Melkonyan, A.A.; Tatikyan, S.Sh.; Capdevielle, J-N

    2002-12-01

    The results of investigations of the neutron component (E=1-10 MeV) of cosmic radiation on board the 'Armenian Airlines' aircrafts using nuclear photoemulsion are presented. The emulsions were exposed on the flights from Yerevan to Moscow, St.-Petersburg, Beirut, Athens, Frankfurt, Amsterdam, Paris and Sofia, and on Concord supersonic flights from Paris to New York. The dependence of the neutron fluxes, and on absorbed and equivalent doses on the flight parameters were investigated. On the flights of the supersonic Concord, with an altitude of 17 km, the neutron fluxes were essentially higher in comparison to those measured on Armenian airliners. It is interesting to note, that the neutron flux and equivalent dose rate decrease with altitude up to 470 km in space, for example, on board the STS-57. The shape of the differential energy spectrum for fast neutrons is the same on all Armenian airlines flights, but significantly different at 17 km altitude, where the flux in the energy region above 3 MeV is increasing.

  4. Investigation of percentage depth dose and dose rate dependence of PAGAT polymer gel dosimeter for photon beams using MRI technique

    International Nuclear Information System (INIS)

    Azadbakht, B.; Hadad, K.; Zahmatkesh, M. H.

    2010-01-01

    In this work, the investigation of the normoxic Pagan polymer-gel dosimeter percentage depth dose and it's dose rate dependence has been made. Using MRI, the formulation to give the maximum change in the transverse relaxation rate R2 was determined to be 4.5% N,N'-methylene-bis-acrylamide(bis), 4.5% acrylamide, 5% gelatine, 5 m M tetrakis (hydroxymethyl) phosphonium chloride, 0.01 m M hydroquinone and 86% HPLC(Water). Irradiation of vials was performed using photon beams of Co-60 therapy unit and an Elec ta linear accelerator. Gel dosimeters were imaged in a Siemens Symphony 1.5 Tesla clinical MRI scanner using a head coil. Post-manufacture irradiation and post imaging times were both selected to be 1 day. For determining the percentage depth dose of the Pagan gel it was found that at the depth of 21 cm, the percentage depth dose for 1.25 MeV γ-ray photons of 60 Co and for 4,6 and 18 MV x-ray photons of Elec ta linear accelerator, are 48%, 52%, 57.3% and 59.73%, respectively. Thus, in the case of the higher energy photon beams, a higher dose can he delivered. to deep-seated tumors. The dose rate dependence of percentage depth dose was studied for 6 MV x-ray photons with the use of dose rates of 80, 160, 240, 320, 400 and 480 c Gy/min. No trend in polymer-gel dosimeter 1/T 2 dependence was found on the mean dose rate and energy for the photon beams.

  5. Understanding differences in dose-equivalents reported by passive and electronic personal dosemeters

    International Nuclear Information System (INIS)

    Perks, Christopher A.; Burgess, Peter; Smith, David; Salasky, Mark; Yahnke, Cliff

    2008-01-01

    Full text: In a number of challenging environments, clients occasionally double badge with electronic personal dosemeters (EPDs) to ensure day-to-day management of their employees personal dose-equivalent while using passive (in our case Luxel or InLight) dosemeters for monthly monitoring for approved results for dose record keeping. In some cases there have been significant differences in the cumulative doses recorded by the EPDs and the passive dosemeters. In these circumstances the passive dosemeters usually report a higher dose than the EPD by up to a factor of two, though more commonly 1.3. In this paper we describe the differences seen between EPD and passive dosemeters (in a number of countries). We then examine the possible causes for such discrepancies by comparison with published response function data available for the EPDs and Landauer dosemeters. We have undertaken a number of experiments comparing directly a number of EPDs and passive dosemeter response to a variety of energy and complex angular geometries where the two types of dosimeter have been exposed at the same time. Recommendations will be made on the appropriate use of double badging in difficult environments and interpretation of the results. (author)

  6. Procedures for operational monitoring of the environmental equivalent doses in cobalt therapy

    International Nuclear Information System (INIS)

    Perez Velasquez, Reytel; Gonzalez Lopez, Nadia; Perez Tamayo, Luis

    2009-01-01

    It took as its object of study environmental equivalent dose rates of field radiation they face radiophysicists technicians and medical physicists in the irradiation site of the radiation department of Lenin Hospital in Holguin, and the public that travels or remains in premises and areas surrounding the campus. It A review of national and international publications as well as technical documents to study the state of the art of the methodological existing monitoring of those dose rates and the valuation of impact in the context of an environmental management system. Since no detailed instructions and to perform the above-mentioned monitoring was proposed structure should contain a procedure to regulate the steps for this monitoring in the radiotherapy department of the Lenin Hospital Holguin, for which we studied the guidelines of NC ISO 14000 , and was conducted wide experiment whose led to: illustrate the level of doses required workers exposed occupationally exposed to radiation and the public compare these levels with natural radiation sources, assess the effectiveness of shielding the site of irradiation and the points of greatest risk within and outside the enclosure irradiation.Also assessed the impact it can have on the health of people exposed to such doses. Finally, we proposed a procedure for conducting subsequent monitoring and made recommendations to reduce levels radiation determined the lowest level reasonably achievable. (author)

  7. Dosimetric evaluation of photon dose calculation under jaw and MLC shielding

    International Nuclear Information System (INIS)

    Fogliata, A.; Clivio, A.; Vanetti, E.; Nicolini, G.; Belosi, M. F.; Cozzi, L.

    2013-01-01

    Purpose: The accuracy of photon dose calculation algorithms in out-of-field regions is often neglected, despite its importance for organs at risk and peripheral dose evaluation. The present work has assessed this for the anisotropic analytical algorithm (AAA) and the Acuros-XB algorithms implemented in the Eclipse treatment planning system. Specifically, the regions shielded by the jaw, or the MLC, or both MLC and jaw for flattened and unflattened beams have been studied.Methods: The accuracy in out-of-field dose under different conditions was studied for two different algorithms. Measured depth doses out of the field, for different field sizes and various distances from the beam edge were compared with the corresponding AAA and Acuros-XB calculations in water. Four volumetric modulated arc therapy plans (in the RapidArc form) were optimized in a water equivalent phantom, PTW Octavius, to obtain a region always shielded by the MLC (or MLC and jaw) during the delivery. Doses to different points located in the shielded region and in a target-like structure were measured with an ion chamber, and results were compared with the AAA and Acuros-XB calculations. Photon beams of 6 and 10 MV, flattened and unflattened were used for the tests.Results: Good agreement between calculated and measured depth doses was found using both algorithms for all points measured at depth greater than 3 cm. The mean dose differences (±1SD) were −8%± 16%, −3%± 15%, −16%± 18%, and −9%± 16% for measurements vs AAA calculations and −10%± 14%, −5%± 12%, −19%± 17%, and −13%± 14% for Acuros-XB, for 6X, 6 flattening-filter free (FFF), 10X, and 10FFF beams, respectively. The same figures for dose differences relative to the open beam central axis dose were: −0.1%± 0.3%, 0.0%± 0.4%, −0.3%± 0.3%, and −0.1%± 0.3% for AAA and −0.2%± 0.4%, −0.1%± 0.4%, −0.5%± 0.5%, and −0.3%± 0.4% for Acuros-XB. Buildup dose was overestimated with AAA, while Acuros-XB gave

  8. Neutron spectrometry and determination of neutron ambient dose equivalents in different LINAC radiotherapy rooms

    International Nuclear Information System (INIS)

    Domingo, C.; Garcia-Fuste, M.J.; Morales, E.; Amgarou, K.; Terron, J.A.; Rosello, J.; Brualla, L.; Nunez, L.; Colmenares, R.; Gomez, F.; Hartmann, G.H.; Sanchez-Doblado, F.; Fernandez, F.

    2010-01-01

    A project has been set up to study the effect on a radiotherapy patient of the neutrons produced around the LINAC accelerator head by photonuclear reactions induced by photons above ∼8 MeV. These neutrons may reach directly the patient, or they may interact with the surrounding materials until they become thermalised, scattering all over the treatment room and affecting the patient as well, contributing to peripheral dose. Spectrometry was performed with a calibrated and validated set of Bonner spheres at a point located at 50 cm from the isocenter, as well as at the place where a digital device for measuring neutrons, based on the upset of SRAM memories induced by thermal neutrons, is located inside the treatment room. Exposures have taken place in six LINAC accelerators with different energies (from 15 to 23 MV) with the aim of relating the spectrometer measurements with the readings of the digital device under various exposure and room geometry conditions. The final purpose of the project is to be able to relate, under any given treatment condition and room geometry, the readings of this digital device to patient neutron effective dose and peripheral dose in organs of interest. This would allow inferring the probability of developing second malignancies as a consequence of the treatment. Results indicate that unit neutron fluence spectra at 50 cm from the isocenter do not depend on accelerator characteristics, while spectra at the place of the digital device are strongly influenced by the treatment room geometry.

  9. The Evaluation of the 0.07 and 3 mm Dose Equivalent with a Portable Beta Spectrometer

    Science.gov (United States)

    Hoshi, Katsuya; Yoshida, Tadayoshi; Tsujimura, Norio; Okada, Kazuhiko

    Beta spectra of various nuclide species were measured using a commercially available compact spectrometer. The shape of the spectra obtained via the spectrometer was almost similar to that of the theoretical spectra. The beta dose equivalent at any depth was obtained as a product of the measured pulse height spectra and the appropriate conversion coefficients of ICRP Publication 74. The dose rates evaluated from the spectra were comparable with the reference dose rates of standard beta calibration sources. In addition, we were able to determine the dose equivalents with a relative error of indication of 10% without the need for complicated correction.

  10. The evaluation of the 0.07 mm and 3 mm dose equivalent with a portable beta spectrometer

    International Nuclear Information System (INIS)

    Hoshi, Katsuya; Yoshida, Tadayoshi; Tsujimura, Norio; Okada, Kazuhiko

    2016-01-01

    Beta spectra of various nuclide species were measured using a commercially available compact spectrometer. The shape of the spectra obtained via the spectrometer was almost similar to that of the theoretical spectra. The beta dose equivalent at any depth was obtained as a product of the measured pulse height spectra and the appropriate conversion coefficients of ICRP Publication 74. The dose rates evaluated from the spectra were comparable with the reference dose rates of standard beta calibration sources. In addition, we were able to determine the dose equivalents with a relative error of indication of 10% without the need for complicated correction. (author)

  11. Equivalent doses of ionizing radiation received by medical staff at a nuclear medicine department

    International Nuclear Information System (INIS)

    Dziuk, E.; Kowalczyk, A.; Siekierzynski, M.; Jazwinski, J.; Chas, J.; Janiak, M.K.; Palijczuk, D.

    2002-01-01

    Aim: Total annual activity of I-131 used for the treatment of thyroid disorders at the Dept.of Nuclear Medicine, Central Clinical Hospital, Military University School of Medicine, in Warsaw, Poland, equal to 190 GBq; at the same time, total activity of Tc-99m utilized at the same Department for diagnostic purposes reached 1 TBq. As estimated from the radiometer readings, in extreme cases the equivalent at a couple of measurement points at this Department may exceed 200 mSv per year. Thus, in the present study we aimed to assess the potential risk of the exposure of medical personnel of the Department to ionizing radiation. Material and Methods: Polymethacrylate cases each housing four thermoluminescent dosimeters were continuously worn for one year by the nurses and doctors with the dosimeters being replaced by the new ones every three months. In addition, cases containing thermoluminescent dosimeters (three dosimeters per case) were placed in 20 different measurement points across the Department and the monitoring of the doses was carried out continuously for more than six years (from May 1995 to March 2002). Based on the quarterly readings of the dosimeters, equivalent doses were calculated for both the members of the personnel and the measurement space points. Results: The doses registered in the patient rooms ranged 5 to 90 mSv x y -1 , in the application room 10 to 15 mSv x y -1 , in the laboratory rooms 1.5 to 30 mSv x y -1 , and in the waiting room 2 to 6 mSv x y -1 ; no increment above the background level was detected in the nurses' station. Accordingly, the annual doses calculated from the dosimeters worn by the staff ranged 0.2 to 1.1 mSv x y -1 ; these latter findings were confirmed by direct readings from individual film dosimeters additionally worn by the staff members. Conclusion: The obtained results indicate that it is unlikely for the personnel of the monitored Nuclear Medicine Department to receive doses of radiation exceeding 40% of the annual

  12. Improved dose localization with dual energy photon irradiation in treatment of lateralized intracranial malignancies

    International Nuclear Information System (INIS)

    Cooley, G.; Gillin, M.T.; Murray, K.J.; Wilson, J.F.; Janjan, N.A.

    1991-01-01

    Dual energy photon irradiation (6 MV and 20 MV) was compared to conventional treatment planning with 6 MV photons in a lateralized intracranial malignancy. Dose volume analysis was performed of both the tumor plus a 2 cm margin (target volume, TV) and normal tissues (NT). Parallel opposed treatment using weightings of 1:1, 1.5:1, and 2:1 were compared for 6 MV photons alone or in combination with 20 MV photons. Uniform treatment of the TV was accomplished within the 60 Gy isodose. Significant differences were observed, however, in NT volumes receiving greater than or equal to 60 Gy and 45-59 Gy. Dual photon energy reduced treatment of NT volumes to greater than or equal to 60 Gy by 13% (177 cm3 vs 204 cm3 in 2:1 weighting) to 70% (147 cm3 vs 498 cm3 in 1:1 weighting) for comparable plans. Dose optimization was also performed for both 6 MV alone or in combination with 20 MV photons. Usual approaches to achieve dose lateralization with conventional isocentric techniques were applied including parallel opposed 6 MV photons ipsilaterally weighted 3.4:1 (POP), and a 110 degrees arc rotational field used to limit treatment to the eye (ARC). Dual energy photon optimized plans included a three beam parallel opposed plan (TOP) and a mixed photon ipsilateral (IPSI) approach. The technique using parallel opposed 20 MV photons and ipsilateral 6 MV photons (TOP) used beam weightings of 1.1 (contralateral 20 MVX): 1.6 (ipsilateral 6 MVX): 1 (ipsilateral 20 MVX) to achieve dose optimization. The ipsilateral approach with 6 MVX and 20 MVX (IPSI) used beam weightings of 1:1.4, respectively. All optimized plans demonstrated a 41% (120 cm3; POP) to 53% (95 cm3; TOP) improvement over parallel opposed 6 MV photons weighted 2:1 (204 cm3) in NT volume receiving greater than or equal to 60 Gy

  13. Evaluation of ambient dose equivalent rates influenced by vertical and horizontal distribution of radioactive cesium in soil in Fukushima Prefecture

    OpenAIRE

    Malins, Alex; Kurikami, Hiroshi; Nakama, Shigeo; Saito, Tatsuo; Okumura, Masahiko; Machida, Masahiko; Kitamura, Akihiro

    2015-01-01

    The air dose rate in an environment contaminated with 134Cs and 137Cs depends on the amount, depth profile and horizontal distribution of these contaminants within the ground. This paper introduces and verifies a tool that models these variables and calculates ambient dose equivalent rates at 1 m above the ground. Good correlation is found between predicted dose rates and dose rates measured with survey meters in Fukushima Prefecture in areas contaminated with radiocesium from the Fukushima D...

  14. The equivalent doses of indoor radon in some dwellings and enclosed areas in Morocco

    International Nuclear Information System (INIS)

    Hakam, O.; Choukri, J.; Reyss, L.

    2008-01-01

    Full text: The principal source of exposure to radiation for public in built-up areas is known to be the inhalation for radon its short-lived daughters.Most of this exposure occurs inside homes,where many hours are spent each day and where the volumic activity of radon is usually higher than outdoors. The compelling effects of radon and its short-lived decay products spread slowly but surely through a wide range of biological problems encountered in such areas as the mortality rates and lung cancer in uranium mines,the results of experimental work with animals, and the discovery of unsually high levels of radon in the living environments of the general population. As a way of prevention, we have measured the volumic activities of indoor radon-222 and we have calculated their effective equivalent dose in some dwellings and enclosed areas in Morocco. The obtained results show that the effective equivalent dose of activities measured indoor dwellings are inferior to the admissible annual limit fixed by ICRP for population, except in two twons situated in regions rich in phosphate deposits where the calculated doses are slightly upper than this limit. The results obtained for enclosed areas are inferior to the admissible annual limit fixed by ICRP for workers, except in the cave of geophysical observatory situated at depth of-12 meters where the obtained value don't present in risk for workers health because workers pass only a few minutes by day in this cave. The risks related to the volumic activities for indoor radon could be avoided by simple precautions such the continuous ventilation

  15. Disturbance from Am-241 Photons of the Cellular Dose by Am-241 Alpha Emissions: Am-241 as an alternative source of alpha particles to radon daughters

    International Nuclear Information System (INIS)

    Lee, Ki-Man; Kim, Eun-Hee

    2015-01-01

    The Radiation Bioengineering Laboratory (RadBio Lab) at Seoul National University (SNU) has built an Am-241 alpha particle irradiator for study of cellular responses to radiation from radon daughters. The radon daughters of concern that cause internal exposure from inhalation of radon-contaminated air are Po-218, Po-214 and Po-210. In their alpha decay schemes, the yields of photon emissions are negligible. Unfortunately, Am-241, the source of alpha irradiator in RadBio Lab, emits photons at every alpha decay while transforming to Np-237 of long half-life. Employing Am-241 as the source simulating radon daughters, therefore, requires that photon emissions from Am-241 be specified in term of dose contribution. In this study, Monte Carlo calculations have been made to characterize dose contributions of Am-241 photon emissions. This study confirms that disturbance from Am-241 photon emissions of the cellular dose by Am-241 alpha emissions is negligible. Dose contamination fraction from photon emissions was 8.02 .. 10 -6 at 25 mm SSD at maximum. Also, note that LET in tissue-equivalent medium varies within about 20% for alpha particles at energies over 5 MeV

  16. Comparison of adult and child radiation equivalent doses from 2 dental cone-beam computed tomography units.

    Science.gov (United States)

    Al Najjar, Anas; Colosi, Dan; Dauer, Lawrence T; Prins, Robert; Patchell, Gayle; Branets, Iryna; Goren, Arthur D; Faber, Richard D

    2013-06-01

    With the advent of cone-beam computed tomography (CBCT) scans, there has been a transition toward these scans' replacing traditional radiographs for orthodontic diagnosis and treatment planning. Children represent a significant proportion of orthodontic patients. Similar CBCT exposure settings are predicted to result in higher equivalent doses to the head and neck organs in children than in adults. The purpose of this study was to measure the difference in equivalent organ doses from different scanners under similar settings in children compared with adults. Two phantom heads were used, representing a 33-year-old woman and a 5-year-old boy. Optically stimulated dosimeters were placed at 8 key head and neck organs, and equivalent doses to these organs were calculated after scanning. The manufacturers' predefined exposure settings were used. One scanner had a pediatric preset option; the other did not. Scanning the child's phantom head with the adult settings resulted in significantly higher equivalent radiation doses to children compared with adults, ranging from a 117% average ratio of equivalent dose to 341%. Readings at the cervical spine level were decreased significantly, down to 30% of the adult equivalent dose. When the pediatric preset was used for the scans, there was a decrease in the ratio of equivalent dose to the child mandible and thyroid. CBCT scans with adult settings on both phantom heads resulted in higher radiation doses to the head and neck organs in the child compared with the adult. In practice, this might result in excessive radiation to children scanned with default adult settings. Collimation should be used when possible to reduce the radiation dose to the patient. While CBCT scans offer a valuable tool, use of CBCT scans should be justified on a specific case-by-case basis. Copyright © 2013 American Association of Orthodontists. Published by Mosby, Inc. All rights reserved.

  17. Development of a fibre-optic dosemeter to measure the skin dose and percentage depth dose in the build-up region of therapeutic photon beams

    International Nuclear Information System (INIS)

    Kim, K. A.; Yoo, W. J.; Jang, K. W.; Moon, J.; Han, K. T.; Jeon, D.; Park, J. Y.; Cha, E. J.; Lee, B.

    2013-01-01

    In this study, a fibre-optic dosemeter (FOD) using an organic scintillator with a diameter of 0.5 mm for photon-beam therapy dosimetry was fabricated. The fabricated dosemeter has many advantages, including water equivalence, high spatial resolution, remote sensing and real-time measurement. The scintillating light generated from an organic-dosemeter probe embedded in a solid-water stack phantom is guided to a photomultiplier tube and an electrometer via 20 m of plastic optical fibre. Using this FOD, the skin dose and the percentage depth dose in the build-up region according to the depths of a solid-water stack phantom are measured with 6- and 15-MV photon-beam energies with field sizes of 10310 and 20320 cm 2 , respectively. The results are compared with those measured using conventional dosimetry films. It is expected that the proposed FOD can be effectively used in radiotherapy dosimetry for accurate measurement of the skin dose and the depth dose distribution in the build-up region due to its high spatial resolution. (authors)

  18. A novel method for patient exit and entrance dose prediction based on water equivalent path length measured with an amorphous silicon electronic portal imaging device

    Science.gov (United States)

    Kavuma, Awusi; Glegg, Martin; Metwaly, Mohamed; Currie, Garry; Elliott, Alex

    2010-01-01

    In vivo dosimetry is one of the quality assurance tools used in radiotherapy to monitor the dose delivered to the patient. Electronic portal imaging device (EPID) images for a set of solid water phantoms of varying thicknesses were acquired and the data fitted onto a quadratic equation, which relates the reduction in photon beam intensity to the attenuation coefficient and material thickness at a reference condition. The quadratic model is used to convert the measured grey scale value into water equivalent path length (EPL) at each pixel for any material imaged by the detector. For any other non-reference conditions, scatter, field size and MU variation effects on the image were corrected by relative measurements using an ionization chamber and an EPID. The 2D EPL is linked to the percentage exit dose table, for different thicknesses and field sizes, thereby converting the plane pixel values at each point into a 2D dose map. The off-axis ratio is corrected using envelope and boundary profiles generated from the treatment planning system (TPS). The method requires field size, monitor unit and source-to-surface distance (SSD) as clinical input parameters to predict the exit dose, which is then used to determine the entrance dose. The measured pixel dose maps were compared with calculated doses from TPS for both entrance and exit depth of phantom. The gamma index at 3% dose difference (DD) and 3 mm distance to agreement (DTA) resulted in an average of 97% passing for the square fields of 5, 10, 15 and 20 cm. The exit dose EPID dose distributions predicted by the algorithm were in better agreement with TPS-calculated doses than phantom entrance dose distributions.

  19. Evaluation of equivalent doses in {sup 18}F PET/CT using the Monte Carlo method with MCNPX code; Avaliação de doses equivalentes em PET/CT com {sup 18}F utilizando o Método Monte Carlo com código MCNPX

    Energy Technology Data Exchange (ETDEWEB)

    Belinato, Walmir [Instituto Federal de Bahia (IFBA), Vitória da Conquista, BA (Brazil); Santos, William Souza; Perini, Ana Paula; Neves, Lucio Pereira [Universidade Federal de Uberlândia (UFU), Uberlândia, MG (Brazil). Instituto de Física; Caldas, Linda V. E. [Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN/SP), São Paulo, SP (Brazil); Souza, Divanizia N. [Universidade Federal de Sergipe (UFS), São Cristóvão, SE (Brazil)

    2017-07-01

    The present work used the Monte Carlo method (MMC), specifically the Monte Carlo NParticle - MCNPX, to simulate the interaction of radiation involving photons and particles, such as positrons and electrons, with virtual adult anthropomorphic simulators on PET / CT scans and to determine absorbed and equivalent doses in adult male and female patients.

  20. Focusing the photon dose in a head phantom

    International Nuclear Information System (INIS)

    Arkuszewski, J.

    1997-01-01

    The 30 keV photons stereotactic focusing is being examined for two kinds of collimated sources and it has been found less advantageous as compared to the gamma-knife MeV range. (author) 2 figs., 2 refs

  1. Measurement of dose equivalent distribution on-board commercial jet aircraft

    International Nuclear Information System (INIS)

    Kubancak, J.; Ambrozova, I.; Ploc, O.; Pachnerova Brabcova, K.; Stepan, V.; Uchihori, Y.

    2014-01-01

    The annual effective doses of aircrew members often exceed the limit of 1 mSv for the public due to the increased level of cosmic radiation at the flight altitudes, and thus, it is recommended to monitor them [International Commission on Radiation Protection. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60. Ann. ICRP 21(1-3), (1991)]. According to the Monte Carlo simulations [Battistoni, G., Ferrari, A., Pelliccioni, M. and Villari, R. Evaluation of the doses to aircrew members taking into consideration the aircraft structures. Adv. Space Res. 36, 1645-1652 (2005) and Ferrari, A., Pelliccioni, M. and Villari, R. Evaluation of the influence of aircraft shielding on the aircrew exposure through an aircraft mathematical model. Radiat. Prot. Dosim. 108(2), 91-105 (2004)], the ambient dose equivalent rate H*(10) depends on the location in the aircraft. The aim of this article is to experimentally evaluate H*(10) on-board selected types of aircraft. The authors found that H*(10) values are higher in the front and the back of the cabin and lesser in the middle of the cabin. Moreover, total dosimetry characteristics obtained in this way are in a reasonable agreement with other data, in particular with the above-mentioned simulations. (authors)

  2. Intercomparison of personal dose equivalent measurements by active personal dosimeters. Final report of a joint IAEA-EURADOS project

    International Nuclear Information System (INIS)

    2007-11-01

    standard laboratory. The final results, as assessed by the irradiation laboratories and discussed with the APD suppliers, were: - The general dosimetric performance of the tested APD is comparable with the performance of standard passive dosimetric systems; - The accuracy at reference photon radiation, the reproducibility and the repeatability of measurements are even better than for most passive dosimeters; - Only three devices have given satisfactory results both for 60 kV (RQR4) and for 120 kV (RQR9) pulsed radiation. Not all the devices have been designed for any radiation field and the end-user should at least take into account information about the dose equivalent rate and energy ranges before using the dosimeters. The performance results confirm that the IEC standard requirements are adequate but that they can be insufficient for some applications such as with pulsed radiation fields

  3. A study on pre-heat conditions in equivalent-dose estimation of holocene loess using single-aliquot regenerative-dose (SAR) protocol

    International Nuclear Information System (INIS)

    Jia Yaofeng; Huang Chunchang; Pang Jiangli; Lu Xinwei; Zhang Xu

    2007-01-01

    Through various arrangements of pre-heat and cut-heat temperatures in the equivalent-dose estimation of Holocene loess using a Double-SAR dating protocol, the paper estimated the equivalent-doses from several loess samples by application of IRSL and Post-IR OSL signals, respectively. The measured results present that the equivalent-dose depends on the heat temperature, especially depends on the cut-heat temperature, showing the equivalent-dose increases with the cut-heat temperature; a plateau of equivalent-dose appears at the 200-300 degree C preheat temperatures and the 200-240 degree C cut-heat temperatures, furthermore, the equivalent-doses estimated by IRSL and Post-IR OSL signals respectively are close to each other, which resulted from the similar sensitivity change directions of optical stimulated signals and their smaller change ranges in the measurement cycles using the various temperatures of pre-heat and cut-heat. This suggests that the 200-300 degree C pre-heat temperatures and the 200-240 degree C cut-heat temperatures are fit for dating young Holocene loess samples. (authors)

  4. Determining clinical photon beam spectra from measured depth dose with the Cimmino algorithm

    International Nuclear Information System (INIS)

    Bloch, P.; Altschuler, M.D.; Bjaerngard, B.E.; Kassaee, A.; McDonough, J.

    2000-01-01

    A method to determine the spectrum of a clinical photon beam from measured depth-dose data is described. At shallow depths, where the range of Compton-generated electrons increases rapidly with photon energy, the depth dose provides the information to discriminate the spectral contributions. To minimize the influence of contaminating electrons, small (6x6cm2 ) fields were used. The measured depth dose is represented as a linear combination of basis functions, namely the depth doses of monoenergetic photon beams derived by Monte Carlo simulations. The weights of the basis functions were obtained with the Cimmino feasibility algorithm, which examines in each iteration the discrepancy between predicted and measured depth dose. For 6 and 15 MV photon beams of a clinical accelerator, the depth dose obtained from the derived spectral weights was within about 1% of the measured depth dose at all depths. Because the problem is ill conditioned, solutions for the spectrum can fluctuate with energy. Physically realistic smooth spectra for these photon beams appeared when a small margin (about ±1%) was attributed to the measured depth dose. The maximum energy of both derived spectra agreed with the measured energy of the electrons striking the target to within 1 MeV. The use of a feasibility method on minimally relaxed constraints provides realistic spectra quickly and interactively. (author)

  5. Dose-equivalent response CR-39 track detector for personnel neutron dosimetry

    International Nuclear Information System (INIS)

    Oda, K.; Ito, M.; Yoneda, H.; Miyake, H.; Yamamoto, J.; Tsuruta, T.

    1991-01-01

    A dose-equivalent response detector based on CR-39 has been designed to be applied for personnel neutron dosimetry. The intrinsic detection efficiency of bare CR-39 was first evaluated from irradiation experiments with monoenergetic neutrons and theoretical calculations. In the second step, the radiator effect was investigated for the purpose of sensitization to fast neutrons. A two-layer radiator consisting of deuterized dotriacontane (C 32 D 66 ) and polyethylene (CH 2 ) was designed. Finally, we made the CR-39 detector sensitive to thermal neutrons by doping with orthocarbone (B 10 H 12 C 2 ), and also estimated the contribution of albedo neutrons. It was found that the new detector - boron-doped CR-39 with the two-layer radiator - would have a flat response with an error of about 70% in a wide energy region, ranging from thermal to 15 MeV. (orig.)

  6. Effective Dose Equivalent To The Cypriot Population Due To Natural Radiation

    Energy Technology Data Exchange (ETDEWEB)

    Christofides, S [Medical Physics Department, Nicosia General Hospital (Cyprus)

    1994-12-31

    A study was initiated by the Biomedical Research Foundation, two years ago, to estimate the various natural radiation components that contribute to the Effective Dose Equivalent (EDE) to the Cypriot population. The present study has shown that the contribution due to cosmic radiation is estimated to be less than 270 microSiverts per annum, while that due to airborne Rn-222 concentration in Cypriot houses is estimated to be less then 330 microSieverts per annum. The contribution due to terrestrial gamma radiations, which is currently under investigation, is so far estimated to be around 108 microSieverts per annum. Therefore the EDE to the Cypriot population due to natural radiation is likely to be around 700 microSieverts per annum, not taking into account the internal exposure due to other naturally occuring radionuclides. (author). 7 refs, 4 figs, 4 tabs.

  7. Evaluation of energy responses for neutron dose-equivalent meters made in Japan

    International Nuclear Information System (INIS)

    Saegusa, J.; Yoshizawa, M.; Tanimura, Y.; Yoshida, M.; Yamano, T.; Nakaoka, H.

    2004-01-01

    Energy responses of three types of Japanese neutron dose-equivalent (DE) meters were evaluated by Monte Carlo simulations and measurements. The energy responses were evaluated for thermal neutrons, monoenergetic neutrons with energies up to 15.2 MeV, and also for neutrons from such radionuclide sources as 252 Cf and 241 Am-Be. The calculated results were corroborated with the measured ones. The angular dependence of the response and the DE response were also evaluated. As a result, reliable energy responses were obtained by careful simulations of the proportional counter, moderator and absorber of the DE meters. Furthermore, the relationship between pressure of counting gas and response of the DE meter was discussed. By using the obtained responses, relations between predicted readings of the DE meters and true DE values were studied for various workplace spectra

  8. Photon activation therapy with 127I-deoxyuridine: measurement of dose enhancement in cultured mammalian cells

    International Nuclear Information System (INIS)

    Fairchild, R.G.; Laster, B.H.; Commerford, S.L.; Furcinitti, P.S.; Sylvester, B.; Gabel, D.; Popenoe, E.; Foster, S.

    1985-01-01

    A technique for radiation enhancement of conventional photon radiotherapy is outlined which has been called photon activation therapy (PAT) (6). High linear energy transfer (LET) radiations in the form of Auger electron distributions are generated by photons of appropriate energies, through photon activation of stable iodine incorporated as an analog of thymidine (Tyd) in DNA. Of the several halogenated deoxyribonucleosides evaluated, iodinated deoxyuridine (IdUrd) has been chosen as the only Tyd analog providing effective photon activation. This mechanism is combined with radiation sensitization produced by IdUrd to produce an overall radiation enhancement. Calculations show that at 5% replacement (IdUrd for Tyd) therapeutic (TG) will vary from ∼2 (single acute dose) to ∼17 (low dose rates associated with permanent implant brachytherapy). Parameters used in the calculation of TG have been evaluated in cell culture; dose enhancements obtained with x-rays (including photon activation) were found to be significantly higher than values measured with γ-rays (no photon activation). Comparison is made between theoretical and measured values. Because of the evident lack of repair of damage produced by both sensitization and photon activation, significant gains are expected to be realized following protracted irradiations. Exchanges (IdUrd for Tyd) for 105 have been obtained in vivo (murine tumors). The authors believe that the application of PAT would be most advantageous in the treatment of brain tumors (grade IV astrocytomas) with implanted 145 Sm sources

  9. Dose conversion coefficients calculated using a series of adult Japanese voxel phantoms against external photon exposure

    International Nuclear Information System (INIS)

    Sato, Kaoru; Endo, Akira; Saito, Kimiaki

    2008-10-01

    This report presents a complete set of conversion coefficients of organ doses and effective doses calculated for external photon exposure using five Japanese adult voxel phantoms developed at the Japan Atomic Energy Agency (JAEA). At the JAEA, high-resolution Japanese voxel phantoms have been developed to clarify the variation of organ doses due to the anatomical characteristics of Japanese, and three male phantoms (JM, JM2 and Otoko) and two female phantoms (JF and Onago) have been constructed up to now. The conversion coefficients of organ doses and effective doses for the five voxel phantoms have been calculated for six kinds of idealized irradiation geometries from monoenergetic photons ranging from 0.01 to 10 MeV using EGS4, a Monte Carlo code for the simulation of coupled electron-photon transport. The dose conversion coefficients are given as absorbed dose and effective dose per unit air-kerma free-in-air, and are presented in tables and figures. The calculated dose conversion coefficients are compared with those of voxel phantoms based on the Caucasian and the recommended values in ICRP74 in order to discuss (1) variation of organ dose due to the body size and individual anatomy, such as position and shape of organs, and (2) effect of posture on organ doses. The present report provides valuable data to study the influence of the body characteristics of Japanese upon the organ doses and to discuss developing reference Japanese and Asian phantoms. (author)

  10. Numerical absorbed dose distributions inside principal organs of a mathematical anthropomorphic phantom irradiated by monoenergetic photon fields

    International Nuclear Information System (INIS)

    Furstoss, C.; Menard, S.

    2005-01-01

    Full text: Personnel can be exposed to photon or mixed (neutrons and photons) radiations at workplaces for various activities (nuclear fuel cycle, medical sector, research... ). The passive and active personal dosimeters worn on the trunk evaluate the personal dose equivalent Hp(10), defined by ICRP 601 to be an estimator of the effective dose E. However, the angular and energy distributions of the radiations encountered could generate an over or under-estimation of the protection quantity E because of the response of the dosimeters or/and because of the definition of Hp(10) itself. The Institute of Radiological Protection and Nuclear Safety (IRSN) is evaluating the possibility of the measurement of the effective dose E using an instrumented anthropomorphic phantom at workplaces. Such an instrument would allow the control of the suitability of the radiological protection instrumentation used at workplaces for radiation fields which can appreciably differ from the reference ISO radiation fields used to calibrate dosimeters. The objectives of this study are to determine key positions for the future detectors inside and on the phantom, as well as their needed technical characteristics. The simulations of the organ absorbed dose distributions performed using the Monte Carlo code MCNPX2 and the MIRD phantom3 model will allow the determination of the detector locations. This paper will present the first numerical results obtained for monoenergetic parallel photon fields. The effective doses E calculated in an energy range from 15 keV to 10 MeV will be presented and compared with the results of M. Zankl et al., published in the GSF report Bericht 8/974. (author)

  11. Experimental verification of a commercial Monte Carlo-based dose calculation module for high-energy photon beams

    International Nuclear Information System (INIS)

    Kuenzler, Thomas; Fotina, Irina; Stock, Markus; Georg, Dietmar

    2009-01-01

    The dosimetric performance of a Monte Carlo algorithm as implemented in a commercial treatment planning system (iPlan, BrainLAB) was investigated. After commissioning and basic beam data tests in homogenous phantoms, a variety of single regular beams and clinical field arrangements were tested in heterogeneous conditions (conformal therapy, arc therapy and intensity-modulated radiotherapy including simultaneous integrated boosts). More specifically, a cork phantom containing a concave-shaped target was designed to challenge the Monte Carlo algorithm in more complex treatment cases. All test irradiations were performed on an Elekta linac providing 6, 10 and 18 MV photon beams. Absolute and relative dose measurements were performed with ion chambers and near tissue equivalent radiochromic films which were placed within a transverse plane of the cork phantom. For simple fields, a 1D gamma (γ) procedure with a 2% dose difference and a 2 mm distance to agreement (DTA) was applied to depth dose curves, as well as to inplane and crossplane profiles. The average gamma value was 0.21 for all energies of simple test cases. For depth dose curves in asymmetric beams similar gamma results as for symmetric beams were obtained. Simple regular fields showed excellent absolute dosimetric agreement to measurement values with a dose difference of 0.1% ± 0.9% (1 standard deviation) at the dose prescription point. A more detailed analysis at tissue interfaces revealed dose discrepancies of 2.9% for an 18 MV energy 10 x 10 cm 2 field at the first density interface from tissue to lung equivalent material. Small fields (2 x 2 cm 2 ) have their largest discrepancy in the re-build-up at the second interface (from lung to tissue equivalent material), with a local dose difference of about 9% and a DTA of 1.1 mm for 18 MV. Conformal field arrangements, arc therapy, as well as IMRT beams and simultaneous integrated boosts were in good agreement with absolute dose measurements in the

  12. Organ dose conversion coefficients for voxel models of the reference male and female from idealized photon exposures

    International Nuclear Information System (INIS)

    Schlattl, H; Zankl, M; Petoussi-Henss, N

    2007-01-01

    A new series of organ equivalent dose conversion coefficients for whole body external photon exposure is presented for a standardized couple of human voxel models, called Rex and Regina. Irradiations from broad parallel beams in antero-posterior, postero-anterior, left- and right-side lateral directions as well as from a 360 deg. rotational source have been performed numerically by the Monte Carlo transport code EGSnrc. Dose conversion coefficients from an isotropically distributed source were computed, too. The voxel models Rex and Regina originating from real patient CT data comply in body and organ dimensions with the currently valid reference values given by the International Commission on Radiological Protection (ICRP) for the average Caucasian man and woman, respectively. While the equivalent dose conversion coefficients of many organs are in quite good agreement with the reference values of ICRP Publication 74, for some organs and certain geometries the discrepancies amount to 30% or more. Differences between the sexes are of the same order with mostly higher dose conversion coefficients in the smaller female model. However, much smaller deviations from the ICRP values are observed for the resulting effective dose conversion coefficients. With the still valid definition for the effective dose (ICRP Publication 60), the greatest change appears in lateral exposures with a decrease in the new models of at most 9%. However, when the modified definition of the effective dose as suggested by an ICRP draft is applied, the largest deviation from the current reference values is obtained in postero-anterior geometry with a reduction of the effective dose conversion coefficient by at most 12%

  13. Organ dose conversion coefficients for voxel models of the reference male and female from idealized photon exposures

    Science.gov (United States)

    Schlattl, H.; Zankl, M.; Petoussi-Henss, N.

    2007-04-01

    A new series of organ equivalent dose conversion coefficients for whole body external photon exposure is presented for a standardized couple of human voxel models, called Rex and Regina. Irradiations from broad parallel beams in antero-posterior, postero-anterior, left- and right-side lateral directions as well as from a 360° rotational source have been performed numerically by the Monte Carlo transport code EGSnrc. Dose conversion coefficients from an isotropically distributed source were computed, too. The voxel models Rex and Regina originating from real patient CT data comply in body and organ dimensions with the currently valid reference values given by the International Commission on Radiological Protection (ICRP) for the average Caucasian man and woman, respectively. While the equivalent dose conversion coefficients of many organs are in quite good agreement with the reference values of ICRP Publication 74, for some organs and certain geometries the discrepancies amount to 30% or more. Differences between the sexes are of the same order with mostly higher dose conversion coefficients in the smaller female model. However, much smaller deviations from the ICRP values are observed for the resulting effective dose conversion coefficients. With the still valid definition for the effective dose (ICRP Publication 60), the greatest change appears in lateral exposures with a decrease in the new models of at most 9%. However, when the modified definition of the effective dose as suggested by an ICRP draft is applied, the largest deviation from the current reference values is obtained in postero-anterior geometry with a reduction of the effective dose conversion coefficient by at most 12%.

  14. Measured Neutron Spectra and Dose Equivalents From a Mevion Single-Room, Passively Scattered Proton System Used for Craniospinal Irradiation

    Energy Technology Data Exchange (ETDEWEB)

    Howell, Rebecca M., E-mail: rhowell@mdanderson.org [Department of Radiation Physics, The University of Texas M. D. Anderson Cancer Center, Houston, Texas (United States); Burgett, Eric A.; Isaacs, Daniel [Department of Nuclear Engineering, Idaho State University, Pocatello, Idaho (United States); Price Hedrick, Samantha G.; Reilly, Michael P.; Rankine, Leith J.; Grantham, Kevin K.; Perkins, Stephanie; Klein, Eric E. [Department of Radiation Oncology, Washington University, St. Louis, Missouri (United States)

    2016-05-01

    Purpose: To measure, in the setting of typical passively scattered proton craniospinal irradiation (CSI) treatment, the secondary neutron spectra, and use these spectra to calculate dose equivalents for both internal and external neutrons delivered via a Mevion single-room compact proton system. Methods and Materials: Secondary neutron spectra were measured using extended-range Bonner spheres for whole brain, upper spine, and lower spine proton fields. The detector used can discriminate neutrons over the entire range of the energy spectrum encountered in proton therapy. To separately assess internally and externally generated neutrons, each of the fields was delivered with and without a phantom. Average neutron energy, total neutron fluence, and ambient dose equivalent [H* (10)] were calculated for each spectrum. Neutron dose equivalents as a function of depth were estimated by applying published neutron depth–dose data to in-air H* (10) values. Results: For CSI fields, neutron spectra were similar, with a high-energy direct neutron peak, an evaporation peak, a thermal peak, and an intermediate continuum between the evaporation and thermal peaks. Neutrons in the evaporation peak made the largest contribution to dose equivalent. Internal neutrons had a very low to negligible contribution to dose equivalent compared with external neutrons, largely attributed to the measurement location being far outside the primary proton beam. Average energies ranged from 8.6 to 14.5 MeV, whereas fluences ranged from 6.91 × 10{sup 6} to 1.04 × 10{sup 7} n/cm{sup 2}/Gy, and H* (10) ranged from 2.27 to 3.92 mSv/Gy. Conclusions: For CSI treatments delivered with a Mevion single-gantry proton therapy system, we found measured neutron dose was consistent with dose equivalents reported for CSI with other proton beamlines.

  15. Derivation of a reference dose and drinking water equivalent level for 1,2,3-trichloropropane.

    Science.gov (United States)

    Tardiff, Robert G; Carson, M Leigh

    2010-06-01

    In some US potable water supplies, 1,2,3-trichloropropane (TCP) has been present at ranges of non-detect to less than 100 ppb, resulting from past uses. In subchronic oral studies, TCP produced toxicity in kidneys, liver, and other tissues. TCP administered by corn oil gavage in chronic studies produced tumors at multiple sites in rats and mice; however, interpretation of these studies was impeded by substantial premature mortality. Drinking water equivalent levels (DWELs) were estimated for a lifetime of consumption by applying biologically-based safety/risk assessment approaches, including Monte Carlo techniques, and with consideration of kinetics and modes of action, to possibly replace default assumptions. Internationally recognized Frameworks for human relevance of animal data were employed to interpret the findings. Calculated were a reference dose (=39 microg/kg d) for non-cancer and Cancer Values (CV) (=10-14 microg/kg d) based on non-linear dose-response relationships for mutagenicity as a precursor of cancer. Lifetime Average Daily Intakes (LADI) are 3130 and 790-1120 microg/person-d for non-cancer and cancer, respectively. DWELs, estimated by applying a relative source contribution (RSC) of 50% to the LADIs, are 780 and 200-280 microg/L for non-cancer and cancer, respectively. These DWELs may inform establishment of formal/informal guidelines and standards to protect public health. Copyright 2010 Elsevier Ltd. All rights reserved.

  16. The choice of a biological model in assessing internal dose equivalent

    International Nuclear Information System (INIS)

    Parodo, A.; Erre, N.

    1977-01-01

    Many are the biological models related to kinetic behavior of radioactive materials within the organism, or in an organ. This is true particularly for the metabolic kinetics of bone-seekers radionuclides described differently by various authors: as a consequence, different forms of the retention function have been used in calculating internal dose equivalent. In our opinion, the retention functions expressed as linear combinations of exponential terms with negative exponents are preferable. In fact, they can be obtained by coherent compartmental analysis and allow a mathematical formalism fairly well definite and easily adaptable to computers. Moreover, it is possible to make use of graphs and monograms already published. The role of the biological model in internal dosimetry, referred to the reliability of the quantitative informations on the kinetic behavior of the radionuclides in the organism and, therefrom, to the accuracy of the doses calculated, is discussed. By comparing the results obtained with different biological models, one finds that the choice of a model is less important than the choice of the value of the appropriate parameters

  17. Comparison of diagnostic protocols and the equivalent effective dose in renal cancer and herniated lumbar disc

    International Nuclear Information System (INIS)

    Ruiz Perez de Villar, M.J.; Vano Carruana, E.; Lanzos Gonbzalez, E.; Perez Torrubia, A.

    1994-01-01

    Renal cancer (RC) and herniated lumbar disc (HLD) were the two pathologies selected for the study of the diagnostic protocols applied in different centers to determine how their variability is reflected in the effective equivalent dose (EED) and establish the optimal radiological protocol for diagnostic purposes, while using the lowest possible dose. On the basis of 222 case histories, it was observed that the EED resulting from the diagnosis of HLD can vary as much as a factor of 3(6.2-18.9 mSv). Likewise, the EED related to the diagnosis of RC can be modified by a factor of 1.5(32.6-48.3 mSv), depending on the diagnostic protocol employed. It can be considered that the optimal protocol to reach a diagnosis of HLD includes chest x-ray, lumbar spine x-ray and lumbar CT scan, while that required for the diagnosis of RC involves chest x-ray, IVU, abdominal CT scan and digital subtraction angiography. The optimization of the study protocols-especially the reduction of the number of exposures, modernization and quality control of the equipment, among other aspects, can reduce the EED by a factor of 2. (Author)

  18. Optimal use of β-blockers in high-risk hypertension: A guide to dosing equivalence

    Directory of Open Access Journals (Sweden)

    Janet B McGill

    2010-05-01

    Full Text Available Janet B McGillDepartment of Medicine, Washington University School of Medicine, St. Louis, Missouri, USAAbstract: Hypertension is the number one diagnosis made by primary care physicians, placing them in a unique position to prescribe the antihypertensive agent best suited to the individual patient. In individuals with diabetes mellitus, blood pressure (BP levels > 130/80 mmHg confer an even higher risk for cardiovascular and renal disease, and these patients will benefit from aggressive antihypertensive treatment using a combination of agents. β‑blockers are playing an increasingly important role in the management of hypertension in high-risk patients. β‑blockers are a heterogeneous class of agents, and this review presents the differences between β‑blockers and provides evidence-based protocols to assist in understanding dose equivalence in the selection of an optimal regimen in patients with complex needs. The clinical benefits provided by β‑blockers are only effective if patients adhere to medication treatment long term. β‑blockers with proven efficacy, once-daily dosing, and lower side effect profiles may become instrumental in the treatment of hypertensive diabetic and nondiabetic patients.Keywords: antihypertensive, blood pressure, atenolol, carvedilol, labetalol, metoprolol, nebivolol

  19. Investigation of real tissue water equivalent path lengths using an efficient dose extinction method

    Science.gov (United States)

    Zhang, Rongxiao; Baer, Esther; Jee, Kyung-Wook; Sharp, Gregory C.; Flanz, Jay; Lu, Hsiao-Ming

    2017-07-01

    For proton therapy, an accurate conversion of CT HU to relative stopping power (RSP) is essential. Validation of the conversion based on real tissue samples is more direct than the current practice solely based on tissue substitutes and can potentially address variations over the population. Based on a novel dose extinction method, we measured water equivalent path lengths (WEPL) on animal tissue samples to evaluate the accuracy of CT HU to RSP conversion and potential variations over a population. A broad proton beam delivered a spread out Bragg peak to the samples sandwiched between a water tank and a 2D ion-chamber detector. WEPLs of the samples were determined from the transmission dose profiles measured as a function of the water level in the tank. Tissue substitute inserts and Lucite blocks with known WEPLs were used to validate the accuracy. A large number of real tissue samples were measured. Variations of WEPL over different batches of tissue samples were also investigated. The measured WEPLs were compared with those computed from CT scans with the Stoichiometric calibration method. WEPLs were determined within  ±0.5% percentage deviation (% std/mean) and  ±0.5% error for most of the tissue surrogate inserts and the calibration blocks. For biological tissue samples, percentage deviations were within  ±0.3%. No considerable difference (extinction measurement took around 5 min to produce ~1000 WEPL values to be compared with calculations. This dose extinction system measures WEPL efficiently and accurately, which allows the validation of CT HU to RSP conversions based on the WEPL measured for a large number of samples and real tissues.

  20. Dose distribution in lungs and thyroid from scatter photons of x-ray mammography imaging

    International Nuclear Information System (INIS)

    Faghihi, R.; Mehdizadeh, S.

    2006-01-01

    The contribution of scatter photons in dose of mammography image in thyroid and lungs are studied. Thyroid and in the form of distribution function and total delivered dose studied by direct measurement with Thermoluminescence dosimeter. The results of measurements compared to other published measurements and the total dose compared to our modelling with Monte Carlo method.. Our phantoms for direct measurement of Dose are a compressed breast phantom placed on a female RANDO phantom. The results of modelling and measurement are in agreement for the total delivered dose to thyroid and lungs and comparable to doses reported by the other researcher

  1. Surface dose investigation of the flattening filter-free photon beams.

    Science.gov (United States)

    Wang, Yuenan; Khan, Mohammad K; Ting, Joseph Y; Easterling, Stephen B

    2012-06-01

    Flattening filter-free (FFF) x-rays can provide more efficient use of photons and a significant increase of dose rate compared with conventional flattened x-rays, features that are especially beneficial for stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT). The available data on the entrance doses of the FFF photon beams remain limited. The purpose of this study was to investigate the entrance dose of FFF photons in the buildup region and to compare it with that of conventional flattened photons. A Varian TrueBeam linear accelerator has been in full clinical operation with 6-MV and 10-MV FFF and flattened x-ray photons. Entrance dose at the surface was measured using a parallel plate ionization chamber in a solid water phantom with buildup depth = 0~15 mm for 6X and 0~25 mm for 10X. Different field size (FS) patterns were created in the Eclipse Treatment Planning System by multileaf collimator (MLC) rather than jaws (FS = 2 × 2, 3 × 3, 4 × 4, 6 × 6, and 10 × 10 cm(2) by MLC and jaw size = 2.2 × 2.2, 3.2 × 3.2, 4.2 × 4.2, 6 × 6, and 10 × 10 cm(2)). The smallest FS was about four times larger than the ion chamber dimension. All buildup dose measurements were normalized to FS = 10 × 10 cm(2) at the depth of dose maximum (dmax). Good repeatability was demonstrated and surface dose increased linearly with FS for both flattened and FFF photons. The entrance dose of the FFF photons was modestly larger than that of the corresponding flattened photons for both 6X and 10X for different FS ranging from 2 × 2 cm(2) to 10 × 10 cm(2). The FFF photons have a higher entrance dose than that of the corresponding flattened photons for FS smaller than 10 × 10 cm(2). However, the difference is not substantial and may be clinically insignificant. Published by Elsevier Inc.

  2. Surface dose measurements and comparison of unflattened and flattened photon beams

    Directory of Open Access Journals (Sweden)

    Ashokkumar Sigamani

    2016-01-01

    Full Text Available The purpose of this study was to evaluate the central axis dose in the build-up region and the surface dose of a 6 MV and 10 MV flattened photon beam (FB and flattening filter free (FFF therapeutic photon beam for different square field sizes (FSs for a Varian Truebeam linear accelerator using parallel-plate ionization chamber and Gafchromic film. Knowledge of dosimetric characteristics in the build-up region and surface dose of the FFF is essential for clinical care. The dose measurements were also obtained empirically using two different commonly used dosimeters: a p-type photon semiconductor dosimeter and a cylindrical ionization chamber. Surface dose increased linearly with FS for both FB and FFF photon beams. The surface dose values of FFF were higher than the FB FSs. The measured surface dose clearly increases with increasing FS. The FFF beams have a modestly higher surface dose in the build-up region than the FB. The dependence of source to skin distance (SSD is less significant in FFF beams when compared to the flattened beams at extended SSDs.

  3. Eye lens dose correlations with personal dose equivalent and patient exposure in paediatric interventional cardiology performed with a fluoroscopic biplane system.

    Science.gov (United States)

    Alejo, L; Koren, C; Corredoira, E; Sánchez, F; Bayón, J; Serrada, A; Guibelalde, E

    2017-04-01

    To analyse the correlations between the eye lens dose estimates performed with dosimeters placed next to the eyes of paediatric interventional cardiologists working with a biplane system, the personal dose equivalent measured on the thorax and the patient dose. The eye lens dose was estimated in terms of H p (0.07) on a monthly basis, placing optically stimulated luminescence dosimeters (OSLDs) on goggles. The H p (0.07) personal dose equivalent was measured over aprons with whole-body OSLDs. Data on patient dose as recorded by the kerma-area product (P KA ) were collected using an automatic dose management system. The 2 paediatric cardiologists working in the facility were involved in the study, and 222 interventions in a 1-year period were evaluated. The ceiling-suspended screen was often disregarded during interventions. The annual eye lens doses estimated on goggles were 4.13±0.93 and 4.98±1.28mSv. Over the aprons, the doses obtained were 10.83±0.99 and 11.97±1.44mSv. The correlation between the goggles and the apron dose was R 2 =0.89, with a ratio of 0.38. The correlation with the patient dose was R 2 =0.40, with a ratio of 1.79μSvGy -1 cm -2 . The dose per procedure obtained over the aprons was 102±16μSv, and on goggles 40±9μSv. The eye lens dose normalized to P KA was 2.21±0.58μSvGy -1 cm -2 . Measurements of personal dose equivalent over the paediatric cardiologist's apron are useful to estimate eye lens dose levels if no radiation protection devices are typically used. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  4. Calculation of photon dose for Dalat research reactor in case of loss of reactor tank water

    International Nuclear Information System (INIS)

    Le Vinh Vinh; Huynh Ton Nghiem; Nguyen Kien Cuong

    2007-01-01

    Photon sources of actinides and fission products were estimated by ORIGEN2 code with the modified cross-section library for Dalat research reactor (DRR) using new cross-section generated by WIMS-ANL code. Photon sources of reactor tank water calculated from the experimental data. MCNP4C2 with available non-analog Monte Carlo model and ANSI/ANL-6.1.1-1977 flux-to-dose factors were used for dose estimation. The agreement between calculation results and those of measurements showed that the methods and models used to get photon sources and dose were acceptable. In case the reactor water totally leaks out from the reactor tank, the calculated dose is very high at the top of reactor tank while still low in control room. In the reactor hall, the operation staffs can access for emergency works but with time limits. (author)

  5. Absorbed dose calibration factors for parallel-plate chambers in high energy photon beams

    International Nuclear Information System (INIS)

    McEwen, M.R.; Duane, S.; Thomas, R.A.S.

    2002-01-01

    An investigation was carried out into the performance of parallel-plate chambers in 60 Co and MV photon beams. The aim was to derive calibration factors, investigate chamber-to-chamber variability and provide much-needed information on the use of parallel-plate chambers in high-energy X-ray beams. A set of NE2561/NE2611 reference chambers, calibrated against the primary standard graphite calorimeter is used for the dissemination of absorbed dose to water. The parallel-plate chambers were calibrated by comparison with the NPL reference chambers in a water phantom. Two types of parallel-plate chamber were investigated - the NACP -02 and Roos and measurements were made at 60 C0 and 6 linac photon energies (6-19 MV). Calibration factors were derived together with polarity corrections. The standard uncertainty in the calibration of a chamber in terms of absorbed dose to water is estimated to be ±0.75%. The results of the polarity measurements were somewhat confusing. One would expect the correction to be small and previous measurements in electron beams have indicated that there is little variation between chambers of these types. However, some chambers gave unexpectedly large polarity corrections, up to 0.8%. By contrast the measured polarity correction for a NE2611 chamber was less than 0.13% at all energies. The reason for these large polarity corrections is not clear, but experimental error and linac variations have been ruled out. By combining the calibration data for the different chambers it was possible to obtain experimental k Q factors for the two chamber types. It would appear from the data that the variations between chambers of the same type are random and one can therefore define a generic curve for each chamber type. These are presented in Figure 1, together with equivalent data for two cylindrical chamber types - NE2561/NE2611 and NE2571. As can be seen, there is a clear difference between the curves for the cylindrical chambers and those for the

  6. A method, using ICRP 26 weighting factors, to determine effective dose equivalent due to nonuniform external exposures

    International Nuclear Information System (INIS)

    Dyer, S.G.

    1993-01-01

    Westinghouse Savannah River Company (WSRC) has recently implemented a methodology and supporting procedures to calculate effective dose equivalent for external exposures. The calculations are based on ICRP 26 methodology and are used to evaluate exposures when multibadging is used. The methodology is based upon the concept of open-quotes whole bodyclose quotes compartmentalization (i.e., the whole body is separated into seven specific regions of radiological concern and weighted accordingly). The highest dose measured in each compartment is used to determine the weighted dose to that compartment. Benefits of determining effective dose equivalent are compliance with DOE Orders, more accurate dose assessments, and the opportunity for improved worker protection through new ALARA opportunities

  7. Measurements of low photon doses using LiF:Mg,Cu,P and CaF{sub 2}:Cu dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Prokert, K [Dresden Univ. of Technology (Germany). Inst. of Radiation Protection Physics; Mann, G [Dresden Univ. of Technology (Germany). Inst. of Radiation Protection Physics

    1997-03-01

    The new thermoluminophors LiF:Mg, Cu, P and CaF{sub 2}:Cu in form of pellets exhibit a significantly higher TL-response than the well-known dosimeters of the types TLD-100 (LiF:Mg, Ti), TLD-400 (CaF{sub 2}:Mn), TLD-900 (CaSO{sub 4}:Dy), etc. Furthermore, the thermoluminophor LiF:Mg, Cu, P shows besides its high sensitivity a good tissue equivalence and therefore, only a small variation of the dose response with the photon energy. The lower limits of detection of these new materials are about 5 {mu}Gy and 0.2 {mu}Gy resp. Therefore, short term measurements of absorbed dose can be realised in radiation fields at very low dose rates (environmental radiation, scattering radiation at medical equipment`s etc.) with an accuracy of {+-}10%. In the field of environmental monitoring the period of exposure can be limited to about 10 days. Using CaF{sub 2}:Cu detectors an exposure of 24 hours is sufficient for dose measurements with lower accuracy. The reusability of CaF{sub 2}:Cu pellets is guaranteed without loss of sensitivity independently of the application of different reading and annealing procedures. In the case of LiF:Mg, Cu, P detectors special procedures are needed in order to keep constant TL-properties. The results of dose measurements at low dose levels in different radiation fields demonstrate the advantages of these detector types. (orig.)

  8. Magnetic field influences on the lateral dose response functions of photon-beam detectors: MC study of wall-less water-filled detectors with various densities.

    Science.gov (United States)

    Looe, Hui Khee; Delfs, Björn; Poppinga, Daniela; Harder, Dietrich; Poppe, Björn

    2017-06-21

    The distortion of detector reading profiles across photon beams in the presence of magnetic fields is a developing subject of clinical photon-beam dosimetry. The underlying modification by the Lorentz force of a detector's lateral dose response function-the convolution kernel transforming the true cross-beam dose profile in water into the detector reading profile-is here studied for the first time. The three basic convolution kernels, the photon fluence response function, the dose deposition kernel, and the lateral dose response function, of wall-less cylindrical detectors filled with water of low, normal and enhanced density are shown by Monte Carlo simulation to be distorted in the prevailing direction of the Lorentz force. The asymmetric shape changes of these convolution kernels in a water medium and in magnetic fields of up to 1.5 T are confined to the lower millimetre range, and they depend on the photon beam quality, the magnetic flux density and the detector's density. The impact of this distortion on detector reading profiles is demonstrated using a narrow photon beam profile. For clinical applications it appears as favourable that the magnetic flux density dependent distortion of the lateral dose response function, as far as secondary electron transport is concerned, vanishes in the case of water-equivalent detectors of normal water density. By means of secondary electron history backtracing, the spatial distribution of the photon interactions giving rise either directly to secondary electrons or to scattered photons further downstream producing secondary electrons which contribute to the detector's signal, and their lateral shift due to the Lorentz force is elucidated. Electron history backtracing also serves to illustrate the correct treatment of the influences of the Lorentz force in the EGSnrc Monte Carlo code applied in this study.

  9. Three-dimensional photon dose distributions with and without lung corrections for tangential breast intact treatments

    International Nuclear Information System (INIS)

    Chin, L.M.; Cheng, C.W.; Siddon, R.L.; Rice, R.K.; Mijnheer, B.J.; Harris, J.R.

    1989-01-01

    The influence of lung volume and photon energy on the 3-dimensional dose distribution for patients treated by intact breast irradiation is not well established. To investigate this issue, we studied the 3-dimensional dose distributions calculated for an 'average' breast phantom for 60Co, 4 MV, 6 MV, and 8 MV photon beams. For the homogeneous breast, areas of high dose ('hot spots') lie along the periphery of the breast near the posterior plane and near the apex of the breast. The highest dose occurs at the inferior margin of the breast tissue, and this may exceed 125% of the target dose for lower photon energies. The magnitude of these 'hot spots' decreases for higher energy photons. When lung correction is included in the dose calculation, the doses to areas at the left and right margin of the lung volume increase. The magnitude of the increase depends on energy and the patient anatomy. For the 'average' breast phantom (lung density 0.31 g/cm3), the correction factors are between 1.03 to 1.06 depending on the energy used. Higher energy is associated with lower correction factors. Both the ratio-of-TMR and the Batho lung correction methods can predict these corrections within a few percent. The range of depths of the 100% isodose from the skin surface, measured along the perpendicular to the tangent of the skin surface, were also energy dependent. The range was 0.1-0.4 cm for 60Co and 0.5-1.4 cm for 8 MV. We conclude that the use of higher energy photons in the range used here provides lower value of the 'hot spots' compared to lower energy photons, but this needs to be balanced against a possible disadvantage in decreased dose delivered to the skin and superficial portion of the breast

  10. Dose calculation methods in photon beam therapy using energy deposition kernels

    International Nuclear Information System (INIS)

    Ahnesjoe, A.

    1991-01-01

    The problem of calculating accurate dose distributions in treatment planning of megavoltage photon radiation therapy has been studied. New dose calculation algorithms using energy deposition kernels have been developed. The kernels describe the transfer of energy by secondary particles from a primary photon interaction site to its surroundings. Monte Carlo simulations of particle transport have been used for derivation of kernels for primary photon energies form 0.1 MeV to 50 MeV. The trade off between accuracy and calculational speed has been addressed by the development of two algorithms; one point oriented with low computional overhead for interactive use and one for fast and accurate calculation of dose distributions in a 3-dimensional lattice. The latter algorithm models secondary particle transport in heterogeneous tissue by scaling energy deposition kernels with the electron density of the tissue. The accuracy of the methods has been tested using full Monte Carlo simulations for different geometries, and found to be superior to conventional algorithms based on scaling of broad beam dose distributions. Methods have also been developed for characterization of clinical photon beams in entities appropriate for kernel based calculation models. By approximating the spectrum as laterally invariant, an effective spectrum and dose distribution for contaminating charge particles are derived form depth dose distributions measured in water, using analytical constraints. The spectrum is used to calculate kernels by superposition of monoenergetic kernels. The lateral energy fluence distribution is determined by deconvolving measured lateral dose distributions by a corresponding pencil beam kernel. Dose distributions for contaminating photons are described using two different methods, one for estimation of the dose outside of the collimated beam, and the other for calibration of output factors derived from kernel based dose calculations. (au)

  11. Surface dose measurements in and out of field. Implications for breast radiotherapy with megavoltage photon beams

    Energy Technology Data Exchange (ETDEWEB)

    Lonski, Peta; Kron, Tomas [Peter MacCallum Cancer Centre, Melbourne (Australia); RMIT Univ., Melbourne (Australia); Ramachandran, Prabhakar; Franich, Rick [Peter MacCallum Cancer Centre, Melbourne (Australia)

    2017-07-01

    This study examines the difference in surface dose between flat and flattening filter free (FFF) photon beams in the context of breast radiotherapy. The surface dose was measured for 6 MV, 6 MV FFF, 10 MV, 10 MV FFF and 18 MV photon beams using a thin window ionisation chamber for various field sizes. Profiles were acquired to ascertain the change in surface dose off-axis. Out-of-field measurements were included in a clinically representative half beam block tangential breast field. In the field centres of FFF beams the surface dose was found to be increased for small fields and decreased for large fields compared to flat beams. For FFF beams, surface dose was found to decrease off-axis and resulted in lower surface dose out-of-field compared to flat beams.

  12. Assessments of conversion coefficients between equivalent dose and accumulated activity using pre-dose scanning images of patients subjected to radioiodine treatment and the Fax/Egs4 computational model

    International Nuclear Information System (INIS)

    Lopes Filho, Ferdinand de J.; Vieira, Jose W.; Andrade Lima, Fernando R. de

    2008-01-01

    The radioiodine is a technique for treatment of thyroid cancer. In this technique, the patients are submitted to the incorporation of the radioactive substance sodium iodide (Na 131 I), which reacts with physiologically metastasis, thyroid tissue remains of and other organs and tissues of the human body. The locations of these reactions are known as areas of highest concentration, hipercaptured areas, hiperconcentrator areas, 'hot areas' or organ-sources and are viewed through images of nuclear medicine scan known as pre-dose (front and rear). To obtain these images, the patient receives, orally, a quantity of 131 I with low activity (± 74 MBq) and is positioned in the chamber of flicker. According to the attendance of hot areas shown in the images, the doctor determines the nuclear activity to be administered in treatment. This analysis is purely qualitative. In this study, the scanning images of pre-dose were adjusted to the dimensions of FAX voxel phantom, and the hot areas correspond to internal sources of the proposed model. Algorithms were developed to generate particles (photons and electrons) in these regions of the FAX. To estimate the coefficients of conversions between equivalent dose and accumulated activity in major radiosensitive organs, FAX and algorithms source were coupled to the Monte Carlo EGS4 code (Electron Gamma Shower, version 4). With these factors is possible to estimate the equivalent doses in the radiosensitive organs and tissues of patients as long as is know the activity administered and the half-life of organic sources. (author)

  13. A Comparison of Equivalent Doses of Lidocaine and Articaine in Maxillary Posterior Tooth Extractions: Case Series

    Directory of Open Access Journals (Sweden)

    Christopher C. Friedl

    2012-06-01

    Full Text Available Objectives: Local anaesthesia is the standard of care during dental extractions. With the advent of newer local anesthetic agents, it is often difficult for the clinician to decide which agent would be most efficacious in a given clinical scenario. This study assessed the efficacy of equal-milligram doses of lidocaine and articaine in achieving surgical anaesthesia of maxillary posterior teeth diagnosed with irreversible pulpitis. Material and Methods: This case-series evaluated a total of 41 patients diagnosed with irreversible pulpitis in a maxillary posterior tooth. Patients randomly received an infiltration of either 3.6 mL (72 mg 2% lidocaine with 1:100,000 epinephrine or 1.8 mL (72 mg 4% articaine with 1:100,000 epinephrine in the buccal fold and palatal soft tissue adjacent to the tooth. After 10 minutes, initial anaesthesia of the tooth was assessed by introducing a sterile 27-gauge needle into the gingival tissue adjacent to the tooth, followed by relief of the gingival cuff. Successful treatment was considered to have occurred when the tooth was extracted with no reported pain. Data was analyzed with the Fisher’s exact test, unpaired t-test and normality test. Results: Twenty-one patients received lidocaine and 20 received articaine. Forty of the 41 patients achieved initial anaesthesia 10 minutes after injection: 21 after lidocaine and 19 after articaine (P = 0.488. Pain-free extraction was accomplished in 33 patients: 19 after lidocaine and 14 after articaine buccal and palatal infiltrations (P = 0.226. Conclusions: There was no significant difference in efficacy between equivalent doses of lidocaine and articaine in the anaesthesia of maxillary posterior teeth with irreversible pulpitis.

  14. Neutron dose equivalent next to the target shield of a neutron therapy facility using an LET counter

    International Nuclear Information System (INIS)

    Stinchcomb, T.G.; Kuchnir, F.T.

    1981-01-01

    The use of a spherical tissue-equivalent proportional counter for measurements of the lineal energy (y) and derivations of the linear energy transfer (LET) for fast neutrons has the advantage of giving distributions of dose and dose equivalent as functions of either LET or y. A measurement next to the target shielding of the neutron therapy facility at the University of Chicago Hospitals and Clinics (UCHC) is described, and the data processing is outlined. The distributions are presented and compared to those from measurements in the neutron beam. The average quality factors are presented

  15. SOILD: A computer model for calculating the effective dose equivalent from external exposure to distributed gamma sources in soil

    International Nuclear Information System (INIS)

    Chen, S.Y.; LePoire, D.; Yu, C.; Schafetz, S.; Mehta, P.

    1991-01-01

    The SOLID computer model was developed for calculating the effective dose equivalent from external exposure to distributed gamma sources in soil. It is designed to assess external doses under various exposure scenarios that may be encountered in environmental restoration programs. The models four major functional features address (1) dose versus source depth in soil, (2) shielding of clean cover soil, (3) area of contamination, and (4) nonuniform distribution of sources. The model is also capable of adjusting doses when there are variations in soil densities for both source and cover soils. The model is supported by a data base of approximately 500 radionuclides. 4 refs

  16. A survey to establish ambient and personal dose equivalent standards in the X- and γ-ray field

    International Nuclear Information System (INIS)

    Morishita, Yuichiro

    2007-01-01

    National Institute of Advanced Industrial Science and Technology (AIST) develops and supplies standards of ionizing radiations as national primary references. Fundamental matters to establish ambient and personal dose equivalent standards of X- and γ-radiation are reviewed in this report. First, units of radiation dose in measurements of X- and γ-radiation are surveyed. Next, the present status of the preparation of X-radiation standard is explained, and finally the relationship between the physical dose and the radiation-protection dose is described. (J.P.N.)

  17. Equivalence of Gyn GEC-ESTRO guidelines for image guided cervical brachytherapy with EUD-based dose prescription

    International Nuclear Information System (INIS)

    Shaw, William; Rae, William ID; Alber, Markus L

    2013-01-01

    To establish a generalized equivalent uniform dose (gEUD) -based prescription method for Image Guided Brachytherapy (IGBT) that reproduces the Gyn GEC-ESTRO WG (GGE) prescription for cervix carcinoma patients on CT images with limited soft tissue resolution. The equivalence of two IGBT planning approaches was investigated in 20 patients who received external beam radiotherapy (EBT) and 5 concomitant high dose rate IGBT treatments. The GGE planning strategy based on dose to the most exposed 2 cm 3 (D2cc) was used to derive criteria for the gEUD-based planning of the bladder and rectum. The safety of gEUD constraints in terms of GGE criteria was tested by maximizing dose to the gEUD constraints for individual fractions. The gEUD constraints of 3.55 Gy for the rectum and 5.19 Gy for the bladder were derived. Rectum and bladder gEUD-maximized plans resulted in D2cc averages very similar to the initial GGE criteria. Average D2ccs and EUDs from the full treatment course were comparable for the two techniques within both sets of normal tissue constraints. The same was found for the tumor doses. The derived gEUD criteria for normal organs result in GGE-equivalent IGBT treatment plans. The gEUD-based planning considers the entire dose distribution of organs in contrast to a single dose-volume-histogram point

  18. Peripheral dose in photon beams from a linear accelerator with a multileaf collimator

    International Nuclear Information System (INIS)

    Lope Lope, R.; Lozano Flores, F.; Gracia Sorrosal, J.; Font Gomez, J.A.; Hernandez Vitoria, A.

    2001-01-01

    Radiation doses outside the radiotherapy treatment field are of radiation protection interest when anatomical structures with very low dose tolerances might be involved. One of the major sources of peripheral dose, scatter from secondary collimators, depends on the configuration of the collimator. In this study, peripheral dose was measured at two depths for 6 and 18 MV photons from a linac Primus (Siemens) with a multileaf collimator (MLC). Comparative measurements were made both with leaves and with the upper jaw positioned at the field edge near to the detector. Configuring the MLC leaves at the field edge yielded a reduction in peripheral dose. (author)

  19. Development of prediction system of dose equivalent rate around a package

    International Nuclear Information System (INIS)

    Nakao, Tetsuya; Minakami, Goro; Taniuchi, Hiroaki; Fujisawa, Kyosuke; Matsukawa, Yukio; Mimura, Shigemi.

    1993-01-01

    A new system is developed that can evaluate the radiation strength of the source in detail, on the basis of the irradiation history of each fuel assembly in a TN-12 or 12A package, and then to determine the best way to organize the assemblies in the package so that the dose equivalent rate around a package is kept to a minimum. This system for minimizing the danger of radiation for operators involved in packaging and transporting spent fuel was developed for personal computer use, to offer ease in handling and high adaptability. The data input is done in dialogue style, with a variety of check functions. In checks to verify the accuracy of the shielding calculation data in this system by comparing the calculated values with several kinds of measured values, the reliability of this new system has been shown to be very high. Since its high utility has been recognized, the system has already been put into use in actual transportation situations. (J.P.N.)

  20. A computer program to calculate the committed dose equivalent after the inhalation of radioactivity

    International Nuclear Information System (INIS)

    Van der Woude, S.

    1989-03-01

    A growing number of people are, as part of their occupation, at risk of being exposed to radiation originating from sources inside their bodies. The quantification of this exposure is an important part of health physics. The International Commission on Radiological Protection (ICRP) developed a first-order kinetics compartmental model to determine the transport of radioactive material through the human body. The model and the parameters involved in its use, are discussed. A versatile computer program was developed to do the following after the in vivo measurement of either the organ- or whole-body activity: calculate the original amount of radioactive material which was inhaled (intake) by employing the ICRP compartmental model of the human body; compare this intake to calculated reference levels and state any action to be taken for the case under consideration; calculate the committed dose equivalent resulting from this intake. In the execution of the above-mentioned calculations, the computer program makes provision for different aerosol particle sizes and the effect of previous intakes. Model parameters can easily be changed to take the effects of, for instance, medical intervention into account. The computer program and the organization of the data in the input files are such that the computer program can be applied to any first-order kinetics compartmental model. The computer program can also conveniently be used for research on problems related to the application of the ICRP model. 18 refs., 25 figs., 5 tabs

  1. Calculation of the individual effective dose equivalent in Italy following the Chernobyl accident

    International Nuclear Information System (INIS)

    Lotfi, M.; Mancioppi, S.; Piermattei, S.; Tommasino, L.; Azimi-Garakani, D.

    1991-01-01

    Estimates are presented here for the individual effective dose equivalents (EDE) from dietary intake of radiocaesium-contaminated food by the average Italian consumer in different age groups. Food items of consumption rate greater than 50kg/y were included in the food basket studies. In view of the pattern of food consumption in Italy, the radiocaesium concentrations of foodstuffs such as milk, milk products, bread, pasta, meat fruit, vegetables and babyfoods were measured between January 1987 and December 1988 inclusive. These measurements were carried out mainly by the environmental radioactivity laboratories in three administrative districts of Italy. The results show that the total individual EDE values received in 1987 from foodstuffs were around 175, 110 and 70 μSv/year for adults, children (up to 10 years old) and infants (up to one year old), respectively. The corresponding values for 1988 are about 15, 10 and 20 μSv/year, respectively. The EDE committed in the latter half relative to the first half of 1987 was a factor of 2.5 and 3 less for adults and children, respectively, and a factor of about 2 for infants. In 1988, no significant change was observed in EDE values during the year. (author)

  2. Mapping of isoexposure curves for evaluation of equivalent environmental doses for radiodiagnostic mobile equipment

    International Nuclear Information System (INIS)

    Bacelar, Alexandre; Andrade, Jose Rodrigo Mendes; Fischer, Andreia Caroline Fischer da Silveira; Accurso, Andre; Hoff, Gabriela

    2011-01-01

    This paper generates iso exposure curves in areas where the mobile radiodiagnostic equipment are used for evaluation of iso kerma map and the environment equivalent dose (H * (d)). It was used a Shimadzu mobile equipment and two Siemens, with non anthropomorphic scatter. The exposure was measured in a mesh of 4.20 x 4.20 square meter in steps of 30 cm, at half height from the scatterer. The calculation of H * (d) were estimated for a worker present in all the procedures in a period of 11 months, being considered 3.55 m As/examination and 44.5 procedures/month (adult UTI) and 3.16 m As/examination and 20.1 procedure/month (pediatric UTI), and 3.16 m As/examination and 20.1 procedure/month (pediatric UTI). It was observed that there exist points where the H * (d) was over the limit established for the free area inside the radius of 30 cm from the central beam of radiation in the case of pediatric UTI and 60 cm for adult UTI. The points localized 2.1 m from the center presented values lower than 25% of those limit

  3. Effect of Photon Hormesis on Dose Responses to Alpha Particles in Zebrafish Embryos

    Directory of Open Access Journals (Sweden)

    Candy Yuen Ping Ng

    2017-02-01

    Full Text Available Photon hormesis refers to the phenomenon where the biological effect of ionizing radiation with a high linear energy transfer (LET value is diminished by photons with a low LET value. The present paper studied the effect of photon hormesis from X-rays on dose responses to alpha particles using embryos of the zebrafish (Danio rerio as the in vivo vertebrate model. The toxicity of these ionizing radiations in the zebrafish embryos was assessed using the apoptotic counts at 20, 24, or 30 h post fertilization (hpf revealed through acridine orange (AO staining. For alpha-particle doses ≥ 4.4 mGy, the additional X-ray dose of 10 mGy significantly reduced the number of apoptotic cells at 24 hpf, which proved the presence of photon hormesis. Smaller alpha-particle doses might not have inflicted sufficient aggregate damages to trigger photon hormesis. The time gap T between the X-ray (10 mGy and alpha-particle (4.4 mGy exposures was also studied. Photon hormesis was present when T ≤ 30 min, but was absent when T = 60 min, at which time repair of damage induced by alpha particles would have completed to prevent their interactions with those induced by X-rays. Finally, the drop in the apoptotic counts at 24 hpf due to photon hormesis was explained by bringing the apoptotic events earlier to 20 hpf, which strongly supported the removal of aberrant cells through apoptosis as an underlying mechanism for photon hormesis.

  4. Significance and principles of the calculation of the effective dose equivalent for radiological protection of personnel and patients

    International Nuclear Information System (INIS)

    Drexler, G.; Williams, G.

    1985-01-01

    The application of the effective dose equivalent, Hsub(E), concept for radiological protection assessments of occupationally exposed persons is justifiable by the practicability thus achieved with regard to the limiting principles. Nevertheless, it would be proper logic to further use as the basic limiting quantity the real physical dose equivalent of homogeneous whole-body exposure, and for inhomogeneous whole-body irradiation the Hsub(E) value, calculated by means of the concept of the effective dose equivalent. For then the required concepts, models and calculations would not be connected with a basic radiation protection quantity. Application of the effective dose equivalent for radiation protection assessments for patients is misleading and is not practical with regard to assessing an individual or collective radiation risk of patients. The quantity of expected harm would be better suited to this purpose. There is no need to express the radiation risk by a dose quantity, which means careless handling of good information. (orig./WU) [de

  5. Assessment of physician and patient (child and adult) equivalent doses during renal angiography by Monte Carlo method

    International Nuclear Information System (INIS)

    Karimian, A.; Nikparvar, B.; Jabbari, I.

    2014-01-01

    Renal angiography is one of the medical imaging methods in which patient and physician receive high equivalent doses due to long duration of fluoroscopy. In this research, equivalent doses of some radiosensitive tissues of patient (adult and child) and physician during renal angiography have been calculated by using adult and child Oak Ridge National Laboratory phantoms and Monte Carlo method (MCNPX). The results showed, in angiography of right kidney in a child and adult patient, that gall bladder with the amounts of 2.32 and 0.35 mSv, respectively, has received the most equivalent dose. About the physician, left hand, left eye and thymus absorbed the most amounts of doses, means 0.020 mSv. In addition, equivalent doses of the physician's lens eye, thyroid and knees were 0.023, 0.007 and 7.9 - 4 mSv, respectively. Although these values are less than the reported thresholds by ICRP 103, it should be noted that these amounts are related to one examination. (authors)

  6. Simulation and experimental study of an indigenously designed and constructed THGEM-based microdosimeter for dose-equivalent measurement

    International Nuclear Information System (INIS)

    Moslehi, A.; Raisali, G.; Lamehi, M.

    2016-01-01

    Most of the GEM/THGEM-based microdosimetric detectors presented in the literature simulate 2 μm of tissue which results in a flat neutron dose-equivalent response in the MeV region. The objective of this work was to introduce a neutron microdosimeter with a more extended flat response. In this regard, a THGEM-based microdosimeter with plexiglas walls, simulating 1 μm of tissue was designed and constructed. Its performance was investigated by both simulation and experimentation to determine the microdosimetric quantity of “lineal energy”. In the simulation study, lineal energy distribution, mean quality factor and dose-equivalent response of the microdosimeter for eleven neutron energies from 10 keV to 14 MeV, along with the energy spectrum of "2"4"1Am-Be neutrons, were calculated by the Geant4 simulation toolkit. Obtained lineal energy distributions were compatible with the distributions determined by a Rossi counter. Also, the mean quality factors agreed well with the values reported by the ICRU report 40 which confirmed tissue equivalent behavior of the microdosimeter. They were different from the effective quality factor values within 15% between 20 keV and 14 MeV. This led to a flat dose-equivalent response with 20% difference from a median value of 0.82 in the above energy range which was an improvement compared with other THGEM-based detectors, simulating 2 μm of tissue. In spite of the satisfactory determination of the dose-equivalent, the microdosimeter had low detection sensitivity. In the experimental study, the measured lineal energy distribution of "2"4"1Am-Be neutrons was in agreement with the simulated distribution. Further, the measured mean quality factor and dose-equivalent differed by 1.5% and 3.5%, respectively, from the calculated values. Finally, it could be concluded that the investigated microdosimeter reliably determined the desired dose-equivalent value of each neutron field with every energy spectrum lying between 20 keV and

  7. Validating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service

    International Nuclear Information System (INIS)

    Jangda, A.Q.; Hussein, S.

    2012-01-01

    In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World Health Organisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the low-energy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility. (author)

  8. Validating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service.

    Science.gov (United States)

    Jangda, Abdul Qadir; Hussein, Sherali

    2012-05-01

    In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World Health Organisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the low-energy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility.

  9. Calculation by the Monte Carlo method of the equivalent dose received by a human fetus from gamma sources localized in the gastrointestinal tract

    International Nuclear Information System (INIS)

    Segreto, V.S.A.

    1979-01-01

    New uterus positions are proposed and worked out in detail to evaluate the exposure of the human fetus to radiation originated in the gastrointestinal-tract during the pregnancy period. In our evaluation each organ in the gastrointestinal-tract namely stomach, small intestine, transverse colon, ascendent colon, descendent colon, sigmoid and rectum was individually considered. Changes in the position of each of these organs were studied as a function of the uterus growth. There were evaluated cases in which the uterus was in three, six and nine month pregnancy for photon energies of 0.02, 0.05, 0.10, 0.50 and 4 MeV. The average equivalent doses (H) of the uterus, in the uterine wall and in each one of the twelve compartiments which we considered as sub-divisions of the uterus were also determined and discussed. (Auhor) [pt

  10. SU-F-T-431: Dosimetric Validation of Acuros XB Algorithm for Photon Dose Calculation in Water

    Energy Technology Data Exchange (ETDEWEB)

    Kumar, L [Rajiv Gandhi Cancer Institute & Research Center, New Delhi, Delhi (India); Yadav, G; Kishore, V [Bundelkhand Institute of Engineering & Technology, Jhansi, Uttar pradesh (India); Bhushan, M; Samuvel, K; Suhail, M [Rajiv Gandhi Cancer Institute and Research Centre, New Delhi, Delhi (India)

    2016-06-15

    Purpose: To validate the Acuros XB algorithm implemented in Eclipse Treatment planning system version 11 (Varian Medical System, Inc., Palo Alto, CA, USA) for photon dose calculation. Methods: Acuros XB is a Linear Boltzmann transport equation (LBTE) solver that solves LBTE equation explicitly and gives result equivalent to Monte Carlo. 6MV photon beam from Varian Clinac-iX (2300CD) was used for dosimetric validation of Acuros XB. Percentage depth dose (PDD) and profiles (at dmax, 5, 10, 20 and 30 cm) measurements were performed in water for field size ranging from 2×2,4×4, 6×6, 10×10, 20×20, 30×30 and 40×40 cm{sup 2}. Acuros XB results were compared against measurements and anisotropic analytical algorithm (AAA) algorithm. Results: Acuros XB result shows good agreement with measurements, and were comparable to AAA algorithm. Result for PDD and profiles shows less than one percent difference from measurements, and from calculated PDD and profiles by AAA algorithm for all field size. TPS calculated Gamma error histogram values, average gamma errors in PDD curves before dmax and after dmax were 0.28, 0.15 for Acuros XB and 0.24, 0.17 for AAA respectively, average gamma error in profile curves in central region, penumbra region and outside field region were 0.17, 0.21, 0.42 for Acuros XB and 0.10, 0.22, 0.35 for AAA respectively. Conclusion: The dosimetric validation of Acuros XB algorithms in water medium was satisfactory. Acuros XB algorithm has potential to perform photon dose calculation with high accuracy, which is more desirable for modern radiotherapy environment.

  11. Intercomparison of personnel dosimetry for thermal neutron dose equivalent in neutron and gamma-ray mixed fields

    International Nuclear Information System (INIS)

    Ogawa, Yoshihiro

    1985-01-01

    In order to consider the problems concerned with personnel dosimetry using film badges and TLDs, an intercomparison of personnel dosimetry, especially dose equivalent responses of personnel dosimeters to thermal neutron, was carried out in five different neutron and gamma-ray mixed fields at KUR and UTR-KINKI from the practical point of view. For the estimation of thermal neutron dose equivalent, it may be concluded that each personnel dosimeter has good performances in the precision, that is, the standard deviations in the measured values by individual dosimeter were within 24 %, and the dose equivalent responses to thermal neutron were almost independent on cadmium ratio and gamma-ray contamination. However, the relative thermal neutron dose equivalent of individual dosimeter normalized to the ICRP recommended value varied considerably and a difference of about 4 times was observed among the dosimeters. From the results obtained, it is suggested that the standardization of calibration factors and procedures is required from the practical point of radiation protection and safety. (author)

  12. Radon and daughters in cigarette smoke measured with SSNTD and corresponding committed equivalent dose to respiratory tract

    International Nuclear Information System (INIS)

    Misdaq, M.A.; Flata, K.

    2003-01-01

    Uranium ( 238 U) and Thorium ( 232 Th) contents were measured inside various tobacco samples by using a method based on determining detection efficiencies of the CR-39 and LR-115 II solid state nuclear track detector (SSNTD) for the emitted alpha particles. Alpha and beta activities per unit volume, due to radon ( 222 Rn), thoron ( 220 Rn) and their decay products, were evaluated inside cigarette smokes of tobacco samples studied. Annual committed equivalent doses due to short-lived radon decay products from the inhalation of various cigarette smokes were determined in the thoracic and extrathoracic regions of the respiratory tract. Three types of cigarettes made in Morocco of black tobacco show higher annual committed equivalent doses in the extrathoracic and thoracic regions of the respiratory tract than the other studied cigarettes (except one type of cigarettes made in France of yellow tobacco); their corresponding annual committed equivalent dose ratios are larger than 1.8. Measured annual committed equivalent doses ranged from 1.8x10 -9 to 1.10x10 -3 Sv yr -1 in the extrathoracic region and from 1.3x10 -10 to 7.6x10 -6 Sv yr -1 in the thoracic region of the respiratory tract for a smoker consuming 20 cigarettes a day

  13. The distribution of committed dose equivalents to workers exposed to tritium in the luminising industry in the United Kingdom

    International Nuclear Information System (INIS)

    Hipkin, J.

    1977-01-01

    In the United Kingdom tritium has become almost the only radionuclide that is used in luminising. Two distinct methods of luminising are used, one involving the use of tritium gas and the other involving the use of tritium activated luminous paint. All major luminisers have voluntarily taken part in urine monitoring programmes. The analyses have been carried out by the National Radiological Protection Board and estimates of committed dose equivalent have been made from the results. The work presented is an analysis of the committed dose equivalents received by all the individuals monitored in the years 1974, 1975 and 1976. It is shown that doses follow, in general, a lognormal distribution modified only at the high dose end by what must be described as dose management. Further evidence for dose management is seen when the pattern of dose versus time are analysed for selected individuals. It is shown that the maximum permissible dose as recommended by the International Commission on Radiological Protection, is only rarely exceeded. It is also shown that there is a substantial difference in the degree of exposure between workers involved in gaseous tritium luminising and workers using paint luminising. A comparison is made between exposure in gaseous tritium luminising and exposure in another common use of gaseous tritium, ie. the filling of electronic devices with tritium gas. It is shown that exposure is very much less in the electronic device work

  14. Derivation of electron and photon energy spectra from electron beam central axis depth dose curves

    Energy Technology Data Exchange (ETDEWEB)

    Deng Jun [Department of Radiation Oncology, Stanford University School of Medicine, Stanford, CA 94305 (United States)]. E-mail: jun@reyes.stanford.edu; Jiang, Steve B.; Pawlicki, Todd; Li Jinsheng; Ma, C.M. [Department of Radiation Oncology, Stanford University School of Medicine, Stanford, CA 94305 (United States)

    2001-05-01

    A method for deriving the electron and photon energy spectra from electron beam central axis percentage depth dose (PDD) curves has been investigated. The PDD curves of 6, 12 and 20 MeV electron beams obtained from the Monte Carlo full phase space simulations of the Varian linear accelerator treatment head have been used to test the method. We have employed a 'random creep' algorithm to determine the energy spectra of electrons and photons in a clinical electron beam. The fitted electron and photon energy spectra have been compared with the corresponding spectra obtained from the Monte Carlo full phase space simulations. Our fitted energy spectra are in good agreement with the Monte Carlo simulated spectra in terms of peak location, peak width, amplitude and smoothness of the spectrum. In addition, the derived depth dose curves of head-generated photons agree well in both shape and amplitude with those calculated using the full phase space data. The central axis depth dose curves and dose profiles at various depths have been compared using an automated electron beam commissioning procedure. The comparison has demonstrated that our method is capable of deriving the energy spectra for the Varian accelerator electron beams investigated. We have implemented this method in the electron beam commissioning procedure for Monte Carlo electron beam dose calculations. (author)

  15. Simulation studies on a prototype ionization chamber for measurement of personal dose equivalent, Hp(10)

    International Nuclear Information System (INIS)

    Cardoso, J.; Oliveira, C.; Carvalho, A.F.

    2005-01-01

    Full text: The Metrological Laboratory of lonizing Radiation and Radioactivity (LMRIR) of Nuclear and Technological Institute (ITN) has designed and constructed a prototype ionization chamber for direct measurement of the personal dose equivalent, H p (10), similar to the developed by the Physikalisch-Technische Bundesanstalt (PTB) and now commercialized by PTW. Tests already performed had shown that the behaviour of this chamber is very close to the PTB chamber, namely the energy dependence for the x-ray radiation qualities of the ISO 4037-1 narrow series N-30, N-40, N-60, N-80, N-100 and N-120 and also for gamma radiation of 137 Cs and 60 Co. However, the results obtained also show a high dependence on the energy for some incident radiation angles and a low magnitude of the electrical response of the ionization chamber. In order to try to optimize the performance of the chamber, namely to decrease the energy dependence and to improve the magnitude of the electrical response of the ionization chamber, the LMRIR initiated numerical simulation of this ionization chamber using a Monte-Carlo method for simulation of radiation transport using, in a first step, the MCNPX code. So, simulation studies of some physical parameters are been performed in order to optimize the response of the ionization chamber, namely the diameter of the central electrode of the ionization chamber, the thickness of the front wall of the ionization chamber, among others. Preliminary results show that probably the actual geometry of the ionization chamber is not yet the optimized configuration. The simulation study will carry on in order to find the optimum geometry. (author)

  16. Photon dose conversion coefficients for the human teeth in standard irradiation geometries

    Energy Technology Data Exchange (ETDEWEB)

    Ulanovsky, A; Wieser, A; Zankl, M; Jacob, P

    2005-07-01

    Photon dose conversion coefficients for the human tooth materials are computed in energy range from 0.01 to 10 MeV by the Monte Carlo method. The voxel phantom Golem of the human body with newly defined tooth region and a modified version of the EGS4 code have been used to compute the coefficients for 30 tooth cells with different locations and materials. The dose responses are calculated for cells representing buccal and lingual enamel layers. The computed coefficients demonstrate a strong dependence on energy and geometry of the radiation source and a weaker dependence on location of the enamel voxels. For isotropic and rotational radiation fields the enamel dose does not show a significant dependence on tooth sample locations. The computed coefficients are used to convert from absorbed dose in teeth to organ dose or to integral air kerma. Examples of integral conversion factors from enamel dose to air kerma are given for several photon fluences specific for the Mayak reprocessing plant in Russia. The integral conversion factors are strongly affected by the energy and angular distributions of photon fluence, which are important characteristics of an exposure scenario for reconstruction of individual occupational doses. (orig.)

  17. The contribution to surface dose form air scatter in mega voltage photon beams

    International Nuclear Information System (INIS)

    Carolan, M.G.; Butson, M.; Metcalfe, P.

    1996-01-01

    Full text: The minimisation of surface dose is an important requirement in radiotherapy in order to avoid undesirable skin reactions. For this reason significant effort has been expended to avoid and understand photon and electron scatter in the heads of linear accelerators which may contribute to surface dose. In this study we have examined the contribution to surface dose which arises due to scatter in the air above the patient. Experimental investigations of air contributions are difficult to design and execute. Therefore we have used Monte Carlo calculations to determine the effect that the presence of air has on surface dose. Methods: The Los Alamos Monte Carlo Neutron and Photon transport code, MCNP4A which incorporates the ETRAN electron transport code from the Integrated TIGER Series of codes was used for our simulations. The geometry used in the model was a 30 cm cube of water. The dose was tallied in cylindrical elements of 7 cm diameter along the axis of the photon beam. For the first millimetre along the beam axis in the phantom, the dose was determined at 0.1 mm increments in 0.1 mm thick volumes. For depths between 1.0 mm and 15.0 mm the dose was determined every 1 mm in 1 mm cylindrical volumes. This yields a depth dose profile with fine spatial resolution near the phantom surface. Dose was also tallied at depths of 5.0, 10.0, 15.0 and 20.0 cm. The simulations were done assuming a 6 MV photon source with a diameter of 1.5 cm, a gaussian intensity profile and a photon energy spectrum based on Mohan et al. (Med. Phys. 12 (1985) 592). No accelerator head geometry was modelled. The field size was defined by virtual collimators which were simply thin regions of zero photon importance and therefore do not contribute to photon or electron scatter. All simulations were run for sufficient particle histories (∼2x10 7 - 5x10 7 source photons) to give statistical uncertainties of ≤ 10% and in most cases ≤ 5%. Fields of size 10, 15, 20 and 25 cm were used

  18. Out-of-Field Dose Equivalents Delivered by Passively Scattered Therapeutic Proton Beams for Clinically Relevant Field Configurations

    International Nuclear Information System (INIS)

    Wroe, Andrew; Clasie, Ben; Kooy, Hanne; Flanz, Jay; Schulte, Reinhard; Rosenfeld, Anatoly

    2009-01-01

    Purpose: Microdosimetric measurements were performed at Massachusetts General Hospital, Boston, MA, to assess the dose equivalent external to passively delivered proton fields for various clinical treatment scenarios. Methods and Materials: Treatment fields evaluated included a prostate cancer field, cranial and spinal medulloblastoma fields, ocular melanoma field, and a field for an intracranial stereotactic treatment. Measurements were completed with patient-specific configurations of clinically relevant treatment settings using a silicon-on-insulator microdosimeter placed on the surface of and at various depths within a homogeneous Lucite phantom. The dose equivalent and average quality factor were assessed as a function of both lateral displacement from the treatment field edge and distance downstream of the beam's distal edge. Results: Dose-equivalent value range was 8.3-0.3 mSv/Gy (2.5-60-cm lateral displacement) for a typical prostate cancer field, 10.8-0.58 mSv/Gy (2.5-40-cm lateral displacement) for the cranial medulloblastoma field, 2.5-0.58 mSv/Gy (5-20-cm lateral displacement) for the spinal medulloblastoma field, and 0.5-0.08 mSv/Gy (2.5-10-cm lateral displacement) for the ocular melanoma field. Measurements of external field dose equivalent for the stereotactic field case showed differences as high as 50% depending on the modality of beam collimation. Average quality factors derived from this work ranged from 2-7, with the value dependent on the position within the phantom in relation to the primary beam. Conclusions: This work provides a valuable and clinically relevant comparison of the external field dose equivalents for various passively scattered proton treatment fields

  19. Approaches to reducing photon dose calculation errors near metal implants

    Energy Technology Data Exchange (ETDEWEB)

    Huang, Jessie Y.; Followill, David S.; Howell, Rebecca M.; Mirkovic, Dragan; Kry, Stephen F., E-mail: sfkry@mdanderson.org [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 and Graduate School of Biomedical Sciences, The University of Texas Health Science Center Houston, Houston, Texas 77030 (United States); Liu, Xinming [Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 and Graduate School of Biomedical Sciences, The University of Texas Health Science Center Houston, Houston, Texas 77030 (United States); Stingo, Francesco C. [Department of Biostatistics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 and Graduate School of Biomedical Sciences, The University of Texas Health Science Center Houston, Houston, Texas 77030 (United States)

    2016-09-15

    Purpose: Dose calculation errors near metal implants are caused by limitations of the dose calculation algorithm in modeling tissue/metal interface effects as well as density assignment errors caused by imaging artifacts. The purpose of this study was to investigate two strategies for reducing dose calculation errors near metal implants: implementation of metal-based energy deposition kernels in the convolution/superposition (C/S) dose calculation method and use of metal artifact reduction methods for computed tomography (CT) imaging. Methods: Both error reduction strategies were investigated using a simple geometric slab phantom with a rectangular metal insert (composed of titanium or Cerrobend), as well as two anthropomorphic phantoms (one with spinal hardware and one with dental fillings), designed to mimic relevant clinical scenarios. To assess the dosimetric impact of metal kernels, the authors implemented titanium and silver kernels in a commercial collapsed cone C/S algorithm. To assess the impact of CT metal artifact reduction methods, the authors performed dose calculations using baseline imaging techniques (uncorrected 120 kVp imaging) and three commercial metal artifact reduction methods: Philips Healthcare’s O-MAR, GE Healthcare’s monochromatic gemstone spectral imaging (GSI) using dual-energy CT, and GSI with metal artifact reduction software (MARS) applied. For the simple geometric phantom, radiochromic film was used to measure dose upstream and downstream of metal inserts. For the anthropomorphic phantoms, ion chambers and radiochromic film were used to quantify the benefit of the error reduction strategies. Results: Metal kernels did not universally improve accuracy but rather resulted in better accuracy upstream of metal implants and decreased accuracy directly downstream. For the clinical cases (spinal hardware and dental fillings), metal kernels had very little impact on the dose calculation accuracy (<1.0%). Of the commercial CT artifact

  20. Approaches to reducing photon dose calculation errors near metal implants

    International Nuclear Information System (INIS)

    Huang, Jessie Y.; Followill, David S.; Howell, Rebecca M.; Mirkovic, Dragan; Kry, Stephen F.; Liu, Xinming; Stingo, Francesco C.

    2016-01-01

    Purpose: Dose calculation errors near metal implants are caused by limitations of the dose calculation algorithm in modeling tissue/metal interface effects as well as density assignment errors caused by imaging artifacts. The purpose of this study was to investigate two strategies for reducing dose calculation errors near metal implants: implementation of metal-based energy deposition kernels in the convolution/superposition (C/S) dose calculation method and use of metal artifact reduction methods for computed tomography (CT) imaging. Methods: Both error reduction strategies were investigated using a simple geometric slab phantom with a rectangular metal insert (composed of titanium or Cerrobend), as well as two anthropomorphic phantoms (one with spinal hardware and one with dental fillings), designed to mimic relevant clinical scenarios. To assess the dosimetric impact of metal kernels, the authors implemented titanium and silver kernels in a commercial collapsed cone C/S algorithm. To assess the impact of CT metal artifact reduction methods, the authors performed dose calculations using baseline imaging techniques (uncorrected 120 kVp imaging) and three commercial metal artifact reduction methods: Philips Healthcare’s O-MAR, GE Healthcare’s monochromatic gemstone spectral imaging (GSI) using dual-energy CT, and GSI with metal artifact reduction software (MARS) applied. For the simple geometric phantom, radiochromic film was used to measure dose upstream and downstream of metal inserts. For the anthropomorphic phantoms, ion chambers and radiochromic film were used to quantify the benefit of the error reduction strategies. Results: Metal kernels did not universally improve accuracy but rather resulted in better accuracy upstream of metal implants and decreased accuracy directly downstream. For the clinical cases (spinal hardware and dental fillings), metal kernels had very little impact on the dose calculation accuracy (<1.0%). Of the commercial CT artifact

  1. Conversion factors for the ICRU dose equivalent quantities for calibrating radiation dosemeters

    International Nuclear Information System (INIS)

    Grosswendt, B.; Hohlfeld, K.; Kramer, H.M.; Selbach, H.J.

    1985-02-01

    Report describing the application of conversion factors for monoenergetic photon radiation and for X and gamma reference radiation used for dosemeter calibration with the aid of spherical or rectangular phantoms (environmental and individual monitoring). (DG) [de

  2. Total external dose equivalent and effective dose derived to the Piedmont population in the period 30 Apr 1986 - 22 Sep 1987

    International Nuclear Information System (INIS)

    Cortissone, C.; Giacomelli, R.; Spezzano, P.; Porzio, L.

    1988-01-01

    Some dosimetric evaluations concerning the population of the Piemonte Region, obtained after the measurements made following the Chernobyl accident, are reported. The individual effective mean dose equivalent derived from the exposure and intake in the period April 30, 1986 - September 22, 1987, is about 0.50 mSv

  3. Calculation of dose distribution for 252Cf fission neutron source in tissue equivalent phantoms using Monte Carlo method

    International Nuclear Information System (INIS)

    Ji Gang; Guo Yong; Luo Yisheng; Zhang Wenzhong

    2001-01-01

    Objective: To provide useful parameters for neutron radiotherapy, the author presents results of a Monte Carlo simulation study investigating the dosimetric characteristics of linear 252 Cf fission neutron sources. Methods: A 252 Cf fission source and tissue equivalent phantom were modeled. The dose of neutron and gamma radiations were calculated using Monte Carlo Code. Results: The dose of neutron and gamma at several positions for 252 Cf in the phantom made of equivalent materials to water, blood, muscle, skin, bone and lung were calculated. Conclusion: The results by Monte Carlo methods were compared with the data by measurement and references. According to the calculation, the method using water phantom to simulate local tissues such as muscle, blood and skin is reasonable for the calculation and measurements of dose distribution for 252 Cf

  4. Method for measuring and evaluation dose equivalent rate from fast neutrons in mixed gamma-neutron fields around particles accelerators

    International Nuclear Information System (INIS)

    Cruceru, I.; Sandu, M.; Cruceru, M.

    1994-01-01

    A method for measuring and evaluation of doses and dose equivalent rate in mixed gamma- neutron fields is discussed in this paper. The method is basedon a double detector system consist of an ionization chamber with components made from a plastic scintillator, coupled to on photomultiplier. Generally the radiation fields around accelerators are complex, often consisting of many different ionizing radiations extending over a broad range of energies. This method solve two major difficulties: determination of response functions of radiation detectors; interpretation of measurement and determination of accuracy. The discrimination gamma-fast neutrons is assured directly without a pulse shape discrimination circuit. The method is applied to mixed fields in which particle energies are situated in the energy range under 20 MeV and an izotropic emision (Φ=10 4 -10 11 n.s -1 ). The dose equivalent rates explored is 0.01mSV--0.1SV

  5. Calculation of effective dose in whole body in dependence of angle of collimator for photon fields

    International Nuclear Information System (INIS)

    Fuenzalida, M.; Varon, C.; Piriz, G.; Banguero, Y.; Lozano, E.; Mancilla, C.

    2011-01-01

    The objective of this work is to obtain quantifiable data of whole body effective dose for photons fields of 6 MV and 18 MV in function of the collimator angle of a Varian Clinac 21EX lineal accelerator. It has been made a variety of studies which investigate the form to reduce the dose in whole body with photons fields, specially over the potential risks and the influence of the collimator angle, as performed Stanthakis et al. [1] with the Monte Carlo method. As a result of this work, the values of whole body effective doses are higher with a 0 deg collimator than with a 90 deg collimator, and as the field size increases, the effective doses difference in whole body, between 0 deg and 90 deg collimator angle, for both energies, becomes smaller. (author)

  6. Calculation of effective dose in whole body in dependence of angle of collimator for photon fields

    Energy Technology Data Exchange (ETDEWEB)

    Fuenzalida, M. [Universidad de la Frontera, Temuco (Chile). Programa de Magister en Fisica Medica; Varon, C.; Piriz, G.; Banguero, Y.; Lozano, E.; Mancilla, C., E-mail: fisicamedica@incancer.c [Instituto Nacional del Cancer, Santiago (Chile). Unidad de Fisica Medica

    2011-07-01

    The objective of this work is to obtain quantifiable data of whole body effective dose for photons fields of 6 MV and 18 MV in function of the collimator angle of a Varian Clinac 21EX lineal accelerator. It has been made a variety of studies which investigate the form to reduce the dose in whole body with photons fields, specially over the potential risks and the influence of the collimator angle, as performed Stanthakis et al. [1] with the Monte Carlo method. As a result of this work, the values of whole body effective doses are higher with a 0 deg collimator than with a 90 deg collimator, and as the field size increases, the effective doses difference in whole body, between 0 deg and 90 deg collimator angle, for both energies, becomes smaller. (author)

  7. Effect of silicone gel breast prosthesis on electron and photon dose distributions

    International Nuclear Information System (INIS)

    Krishnan, L.; St George, F.J.; Mansfield, C.M.; Krishnan, E.C.

    1983-01-01

    The effect of a silicone gel breast prosthesis on the absorbed dose distribution of 9-20 MeV electron beams and 1.25-15 MV photon beams was studied. Compared to water measurements, at depths smaller than the practical range of the electron beams, the central axis depth dose values below the prosthesis were lower for all energies by as much as 3.5%. However, at depths near the practical range, the central axis depth dose values for the prosthesis were greater than that of water by as much as 33%. Since this occurs near the end of the electron range, the resultant difference may not be clinically significant. Results of the effect of breast prosthesis on photon depth dose distributions reveal that no clinically significant perturbation is produced by the breast prosthesis using Co-60, 6- and 15-MV radiations

  8. Effect of silicone gel breast prosthesis on electron and photon dose distributions

    International Nuclear Information System (INIS)

    Krishnan, L.; St George, F.J.; Mansfield, C.M.; Krishnan, E.C.

    1983-01-01

    The effect of a silicone gel breast prosthesis on the absorbed dose distribution of 9--20 MeV electron beams and 1.25--15 MV photon beams was studied. Compared to water measurements, at depths smaller than the practical range of the electron beams, the central axis depth dose values below the prothesis were lower for all energies by as much as 3.5%. However, at depths near the practical range, the central axis depth dose values for the prosthesis were greater than that of water by as much as 33%. Since this occurs near the end of the electron range, the resultant difference may not be clinically significant. Results of the effect of breast prosthesis on photon depth dose distributions reveal that no clinically significant perturbation is produced by the breast prosthesis using Co-60, 6- and 15-MV radiations

  9. Modelling of electron contamination in clinical photon beams for Monte Carlo dose calculation

    International Nuclear Information System (INIS)

    Yang, J; Li, J S; Qin, L; Xiong, W; Ma, C-M

    2004-01-01

    The purpose of this work is to model electron contamination in clinical photon beams and to commission the source model using measured data for Monte Carlo treatment planning. In this work, a planar source is used to represent the contaminant electrons at a plane above the upper jaws. The source size depends on the dimensions of the field size at the isocentre. The energy spectra of the contaminant electrons are predetermined using Monte Carlo simulations for photon beams from different clinical accelerators. A 'random creep' method is employed to derive the weight of the electron contamination source by matching Monte Carlo calculated monoenergetic photon and electron percent depth-dose (PDD) curves with measured PDD curves. We have integrated this electron contamination source into a previously developed multiple source model and validated the model for photon beams from Siemens PRIMUS accelerators. The EGS4 based Monte Carlo user code BEAM and MCSIM were used for linac head simulation and dose calculation. The Monte Carlo calculated dose distributions were compared with measured data. Our results showed good agreement (less than 2% or 2 mm) for 6, 10 and 18 MV photon beams

  10. Establishing the impact of temporary tissue expanders on electron and photon beam dose distributions.

    Science.gov (United States)

    Asena, A; Kairn, T; Crowe, S B; Trapp, J V

    2015-05-01

    This study investigates the effects of temporary tissue expanders (TTEs) on the dose distributions in breast cancer radiotherapy treatments under a variety of conditions. Using EBT2 radiochromic film, both electron and photon beam dose distribution measurements were made for different phantoms, and beam geometries. This was done to establish a more comprehensive understanding of the implant's perturbation effects under a wider variety of conditions. The magnetic disk present in a tissue expander causes a dose reduction of approximately 20% in a photon tangent treatment and 56% in electron boost fields immediately downstream of the implant. The effects of the silicon elastomer are also much more apparent in an electron beam than a photon beam. Evidently, each component of the TTE attenuates the radiation beam to different degrees. This study has demonstrated that the accuracy of photon and electron treatments of post-mastectomy patients is influenced by the presence of a tissue expander for various beam orientations. The impact of TTEs on dose distributions establishes the importance of an accurately modelled high-density implant in the treatment planning system for post-mastectomy patients. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  11. Determination of the absorbed dose and dose-distribution in water for low- and medium-energetic photons

    International Nuclear Information System (INIS)

    Bultman, J.H.

    1990-05-01

    The methods to determine the absorbed dose to water for low and medium energy photons were studied. Large differences between the results of these methods exists. So, a research proposition has been made to explain these differences. The goal of this research will be the development of a method to determine the absorbed dose below approximately 400 keV with an ionization chamber calibrated at 60 Co gamma radiation. To explain the differences between the set of methods, some causes were proposed, like the influence of the ionisation chamber on the measurement in water. Also, some methods to determine the factors are proposed. (author). 29 refs

  12. The biologically equivalent dose BED - Is the approach for calculation of this factor really a reliable basis?

    International Nuclear Information System (INIS)

    Jensen, J.M.; Zimmermann, J.

    2000-01-01

    To predict the effect on tumours in radiotherapy, especially relating to irreversible effects, but also to realize the retrospective assessment the so called L-Q-model is relied on at present. Internal specific organ parameters, such as α, β, γ, T p , T k , and ρ, as well as external parameters, so as D, d, n, V, and V ref , were used for determination of the biologically equivalent dose BED. While the external parameters are determinable with small deviations, the internal parameters depend on biological varieties and dispersons: In some cases the lowest value is assumed to be Δ=±25%. This margin of error goes on to the biologically equivalent dose by means of the principle of superposition of errors. In some selected cases (lung, kidney, skin, rectum) these margins of error were calculated exemplarily. The input errors especially of the internal parameters cause a mean error Δ on the biologically equivalent dose and a dispersion of the single fraction dose d dependent on the organ taking into consideration, of approximately 8-30%. Hence it follows only a very critical and cautious application of those L-Q-algorithms in expert proceedings, and in radiotherapy more experienced based decisions are recommended, instead of acting only upon simple two-dimensional mechanistic ideas. (orig.) [de

  13. Thermally assisted OSL application for equivalent dose estimation; comparison of multiple equivalent dose values as well as saturation levels determined by luminescence and ESR techniques for a sedimentary sample collected from a fault gouge

    Energy Technology Data Exchange (ETDEWEB)

    Şahiner, Eren, E-mail: sahiner@ankara.edu.tr; Meriç, Niyazi, E-mail: meric@ankara.edu.tr; Polymeris, George S., E-mail: gspolymeris@ankara.edu.tr

    2017-02-01

    Highlights: • Multiple equivalent dose estimations were carried out. • Additive ESR and regenerative luminescence were applied. • Preliminary SAR results employing TA-OSL signal were discussed. • Saturation levels of ESR and luminescence were investigated. • IRSL{sub 175} and SAR TA-OSL stand as very promising for large doses. - Abstract: Equivalent dose estimation (D{sub e}) constitutes the most important part of either trap-charge dating techniques or dosimetry applications. In the present work, multiple, independent equivalent dose estimation approaches were adopted, using both luminescence and ESR techniques; two different minerals were studied, namely quartz as well as feldspathic polymineral samples. The work is divided into three independent parts, depending on the type of signal employed. Firstly, different D{sub e} estimation approaches were carried out on both polymineral and contaminated quartz, using single aliquot regenerative dose protocols employing conventional OSL and IRSL signals, acquired at different temperatures. Secondly, ESR equivalent dose estimations using the additive dose procedure both at room temperature and at 90 K were discussed. Lastly, for the first time in the literature, a single aliquot regenerative protocol employing a thermally assisted OSL signal originating from Very Deep Traps was applied for natural minerals. Rejection criteria such as recycling and recovery ratios are also presented. The SAR protocol, whenever applied, provided with compatible D{sub e} estimations with great accuracy, independent on either the type of mineral or the stimulation temperature. Low temperature ESR signals resulting from Al and Ti centers indicate very large D{sub e} values due to bleaching in-ability, associated with large uncertainty values. Additionally, dose saturation of different approaches was investigated. For the signal arising from Very Deep Traps in quartz saturation is extended almost by one order of magnitude. It is

  14. Assessment of population external irradiation doses with consideration of Rospotrebnadzor bodies equipment for monitoring of photon radiation dose

    Directory of Open Access Journals (Sweden)

    I. P. Stamat

    2016-01-01

    Full Text Available This paper provides review of equipment and methodology for measurement of photon radiation dose; analysis of possible reasons for considerable deviation between the Russian Federation population annual effective external irradiation doses and the relevant average global value. Data on Rospotrebnadzor bodies dosimetry equipment used for measurement of gamma radiation dose are collected and systematized. Over 60 kinds of dosimeters are used for monitoring of population external irradiation doses. Most of dosimeters used in the country have gas-discharge detectors (Geiger-Mueller counters, minor biochemical annunciators, etc. which have higher total values of own background level and of space radiation response than the modern dosimeters with scintillation detectors. This feature of dosimeters is apparently one of most plausible reasons of a bit overstating assessment of population external irradiation doses. The options for specification of population external irradiation doses assessment are: correction of gamma radiation dose measurement results with consideration of dosimeters own background level and space radiation response, introduction of more up-to-date dosimeters with scintillation detectors, etc. The most promising direction of research in verification of population external irradiation doses assessment is account of dosimetry equipment.

  15. Relationship between the generalized equivalent uniform dose formulation and the Poisson statistics-based tumor control probability model

    International Nuclear Information System (INIS)

    Zhou Sumin; Das, Shiva; Wang Zhiheng; Marks, Lawrence B.

    2004-01-01

    The generalized equivalent uniform dose (GEUD) model uses a power-law formalism, where the outcome is related to the dose via a power law. We herein investigate the mathematical compatibility between this GEUD model and the Poisson statistics based tumor control probability (TCP) model. The GEUD and TCP formulations are combined and subjected to a compatibility constraint equation. This compatibility constraint equates tumor control probability from the original heterogeneous target dose distribution to that from the homogeneous dose from the GEUD formalism. It is shown that this constraint equation possesses a unique, analytical closed-form solution which relates radiation dose to the tumor cell survival fraction. It is further demonstrated that, when there is no positive threshold or finite critical dose in the tumor response to radiation, this relationship is not bounded within the realistic cell survival limits of 0%-100%. Thus, the GEUD and TCP formalisms are, in general, mathematically inconsistent. However, when a threshold dose or finite critical dose exists in the tumor response to radiation, there is a unique mathematical solution for the tumor cell survival fraction that allows the GEUD and TCP formalisms to coexist, provided that all portions of the tumor are confined within certain specific dose ranges

  16. Photonic activation of plasminogen induced by low dose UVB.

    Directory of Open Access Journals (Sweden)

    Manuel Correia

    Full Text Available Activation of plasminogen to its active form plasmin is essential for several key mechanisms, including the dissolution of blood clots. Activation occurs naturally via enzymatic proteolysis. We report that activation can be achieved with 280 nm light. A 2.6 fold increase in proteolytic activity was observed after 10 min illumination of human plasminogen. Irradiance levels used are in the same order of magnitude of the UVB solar irradiance. Activation is correlated with light induced disruption of disulphide bridges upon UVB excitation of the aromatic residues and with the formation of photochemical products, e.g. dityrosine and N-formylkynurenine. Most of the protein fold is maintained after 10 min illumination since no major changes are observed in the near-UV CD spectrum. Far-UV CD shows loss of secondary structure after illumination (33.4% signal loss at 206 nm. Thermal unfolding CD studies show that plasminogen retains a native like cooperative transition at ~70 ºC after UV-illumination. We propose that UVB activation of plasminogen occurs upon photo-cleavage of a functional allosteric disulphide bond, Cys737-Cys765, located in the catalytic domain and in van der Waals contact with Trp761 (4.3 Å. Such proximity makes its disruption very likely, which may occur upon electron transfer from excited Trp761. Reduction of Cys737-Cys765 will result in likely conformational changes in the catalytic site. Molecular dynamics simulations reveal that reduction of Cys737-Cys765 in plasminogen leads to an increase of the fluctuations of loop 760-765, the S1-entrance frame located close to the active site. These fluctuations affect the range of solvent exposure of the catalytic triad, particularly of Asp646 and Ser74, which acquire an exposure profile similar to the values in plasmin. The presented photonic mechanism of plasminogen activation has the potential to be used in clinical applications, possibly together with other enzymatic treatments for the

  17. The difference of scoring dose to water or tissues in Monte Carlo dose calculations for low energy brachytherapy photon sources.

    Science.gov (United States)

    Landry, Guillaume; Reniers, Brigitte; Pignol, Jean-Philippe; Beaulieu, Luc; Verhaegen, Frank

    2011-03-01

    The goal of this work is to compare D(m,m) (radiation transported in medium; dose scored in medium) and D(w,m) (radiation transported in medium; dose scored in water) obtained from Monte Carlo (MC) simulations for a subset of human tissues of interest in low energy photon brachytherapy. Using low dose rate seeds and an electronic brachytherapy source (EBS), the authors quantify the large cavity theory conversion factors required. The authors also assess whether ap plying large cavity theory utilizing the sources' initial photon spectra and average photon energy induces errors related to spatial spectral variations. First, ideal spherical geometries were investigated, followed by clinical brachytherapy LDR seed implants for breast and prostate cancer patients. Two types of dose calculations are performed with the GEANT4 MC code. (1) For several human tissues, dose profiles are obtained in spherical geometries centered on four types of low energy brachytherapy sources: 125I, 103Pd, and 131Cs seeds, as well as an EBS operating at 50 kV. Ratios of D(w,m) over D(m,m) are evaluated in the 0-6 cm range. In addition to mean tissue composition, compositions corresponding to one standard deviation from the mean are also studied. (2) Four clinical breast (using 103Pd) and prostate (using 125I) brachytherapy seed implants are considered. MC dose calculations are performed based on postimplant CT scans using prostate and breast tissue compositions. PTV D90 values are compared for D(w,m) and D(m,m). (1) Differences (D(w,m)/D(m,m)-1) of -3% to 70% are observed for the investigated tissues. For a given tissue, D(w,m)/D(m,m) is similar for all sources within 4% and does not vary more than 2% with distance due to very moderate spectral shifts. Variations of tissue composition about the assumed mean composition influence the conversion factors up to 38%. (2) The ratio of D90(w,m) over D90(m,m) for clinical implants matches D(w,m)/D(m,m) at 1 cm from the single point sources, Given

  18. Advanced prostate cancer: the results of a randomized comparative trial of high dose irradiation boosting with conformal protons compared with conventional dose irradiation using photons alone

    Energy Technology Data Exchange (ETDEWEB)

    Shipley, William U; Verhey, Lynn J; Munzenrider, John E; Suit, Herman D; Urie, Marcia M; McManus, Patricia L; Young, Robert H; Shipley, Jenot W; Zietman, Anthony L; Biggs, Peter J; Heney, Niall M; Goitein, Michael

    1995-04-30

    Purpose: Following a thorough Phase I/II study, we evaluated by a Phase III trial high versus conventional dose external beam irradiation as mono-therapy for patients with Stage T3-T4 prostate cancer. Patient outcome following standard dose radiotherapy or following a 12.5% increase in total dose to 75.6 Cobalt Gray Equivalent (CGE) using a conformal perineal proton boost was compared for local tumor control, disease-free survival, and overall survival. Methods and Materials: Stage T3-T4, Nx, N0-2, M0 patients received 50.4 Gy by four-field photons and were randomized to receive either an additional 25.2 CGE by conformal protons (arm 1--the high dose arm, 103 patients, total dose 75.6 CGE) or an additional 16.8 Gy by photons (arm 2--the conventional dose arm, 99 patients, total dose 67.2 Gy). Actuarial overall survival (OS), disease-specific survival (DSS), total recurrence-free survival (TRFS), (clinically free, prostate specific antigen (PSA) less than 4ng/ml and a negative prostate rebiopsy, done in 38 patients without evidence of disease) and local control (digital rectal exam and rebiopsy negative) were evaluated. Results: The protocol completion rate was 90% for arm 1 and 97% for arm 2. With a median follow-up of 61 months (range 3 to 139 months) 135 patients are alive and 67 have died, 20 from causes other than prostate cancer. We found no significant differences in OS, DSS, TRFS or local control between the two arms. Among those completing randomized treatment (93 in arm 1 and 96 in arm 2), the local control at 5 and 8 years for arm 1 is 92% and 77%, respectively and is 80% and 60%, respectively for arm 2 (p = .089) and there are no significant differences in OS, DSS, and TRFS. The local control for the 57 patients with poorly differentiated (Gleason 4 or 5 of 5) tumors at 5 and 8 years for arm 1 is 94% and 84% and is 64% and 19% on arm 2 (p 0.0014). In patients whose digital rectal exam had normalized following treatment and underwent prostate rebiopsy

  19. Advanced prostate cancer: the results of a randomized comparative trial of high dose irradiation boosting with conformal protons compared with conventional dose irradiation using photons alone

    International Nuclear Information System (INIS)

    Shipley, William U.; Verhey, Lynn J.; Munzenrider, John E.; Suit, Herman D.; Urie, Marcia M.; McManus, Patricia L.; Young, Robert H.; Shipley, Jenot W.; Zietman, Anthony L.; Biggs, Peter J.; Heney, Niall M.; Goitein, Michael

    1995-01-01

    Purpose: Following a thorough Phase I/II study, we evaluated by a Phase III trial high versus conventional dose external beam irradiation as mono-therapy for patients with Stage T3-T4 prostate cancer. Patient outcome following standard dose radiotherapy or following a 12.5% increase in total dose to 75.6 Cobalt Gray Equivalent (CGE) using a conformal perineal proton boost was compared for local tumor control, disease-free survival, and overall survival. Methods and Materials: Stage T3-T4, Nx, N0-2, M0 patients received 50.4 Gy by four-field photons and were randomized to receive either an additional 25.2 CGE by conformal protons (arm 1--the high dose arm, 103 patients, total dose 75.6 CGE) or an additional 16.8 Gy by photons (arm 2--the conventional dose arm, 99 patients, total dose 67.2 Gy). Actuarial overall survival (OS), disease-specific survival (DSS), total recurrence-free survival (TRFS), (clinically free, prostate specific antigen (PSA) less than 4ng/ml and a negative prostate rebiopsy, done in 38 patients without evidence of disease) and local control (digital rectal exam and rebiopsy negative) were evaluated. Results: The protocol completion rate was 90% for arm 1 and 97% for arm 2. With a median follow-up of 61 months (range 3 to 139 months) 135 patients are alive and 67 have died, 20 from causes other than prostate cancer. We found no significant differences in OS, DSS, TRFS or local control between the two arms. Among those completing randomized treatment (93 in arm 1 and 96 in arm 2), the local control at 5 and 8 years for arm 1 is 92% and 77%, respectively and is 80% and 60%, respectively for arm 2 (p = .089) and there are no significant differences in OS, DSS, and TRFS. The local control for the 57 patients with poorly differentiated (Gleason 4 or 5 of 5) tumors at 5 and 8 years for arm 1 is 94% and 84% and is 64% and 19% on arm 2 (p 0.0014). In patients whose digital rectal exam had normalized following treatment and underwent prostate rebiopsy

  20. Photon beam dose distributions for patients with implanted temporary tissue expanders

    Science.gov (United States)

    Asena, A.; Kairn, T.; Crowe, S. B.; Trapp, J. V.

    2015-01-01

    This study examines the effects of temporary tissue expanders (TTEs) on the dose distributions of photon beams in breast cancer radiotherapy treatments. EBT2 radiochromic film and ion chamber measurements were taken to quantify the attenuation and backscatter effects of the inhomogeneity. Results illustrate that the internal magnetic port present in a tissue expander causes a dose reduction of approximately 25% in photon tangent fields immediately downstream of the implant. It was also shown that the silicone elastomer shell of the tissue expander reduced the dose to the target volume by as much as 8%. This work demonstrates the importance for an accurately modelled high-density implant in the treatment planning system for post-mastectomy breast cancer patients.

  1. The Dose Estimation Formula Of Photon Radiation To Film Badge Of Kodak Type 2

    International Nuclear Information System (INIS)

    Rohmah, Nur

    2000-01-01

    Study to determine the formula of dose estimation for photon radiation to film badge of Kodak type 2 has been carried out. The irradiation was done by irradiated film badge of Kodak type 2 using photon sources of X-rays machine, 137 Cs and 60 Co. By determining the apparent dose and also the sensitivity values each filters of the calibration curve and the weighting factors of energy dependence curve, the formula of the dose estimation for film badge of Kodak type 2 could be obtained, i.e. H 1cm 2.066761E-02N ADPI-2 + 1.953342N ADAI - 8.946254N ADCu + 24.80611N ADSn/pb

  2. Semi-empirical equivalent field method for dose determination in midline block fields for cobalt - 60 beam

    International Nuclear Information System (INIS)

    Tagoe, S.N.A.; Nani, E.K.; Yarney, J.; Edusa, C.; Quayson-Sackey, K.; Nyamadi, K.M.; Sasu, E.

    2012-01-01

    For teletherapy treatment time calculations, midline block fields are resolved into two fields, but neglecting scattering from other fields, the effective equivalent square field size of the midline block is assumed to the resultant field. Such approach is underestimation, and may be detrimental in achieving the recommended uncertainty of ± 5 % for patient's radiation dose delivery. By comparison, the deviations of effective equivalent square field sizes by calculations and experiments were within 13.2 % for cobalt 60 beams of GWGP80 cobalt 60 teletherapy. Therefore, a modified method incorporating the scatter contributions was adopted to estimate the effective equivalent square field size for midline block field. The measured outputs of radiation beams with the block were compared with outputs of square fields without the blocks (only the block tray) at depths of 5 and 10 cm for the teletherapy machine employing isocentric technique, and the accuracy was within ± 3 % for the cobalt 60 beams. (au)

  3. Medical irradiation and the use of the ''effective dose equivalent'' concept

    International Nuclear Information System (INIS)

    Persson, B.R.R.

    1980-01-01

    The aim of this paper is to demonstrate the use of the effective dose for all kinds of medical irradiation. In order to estimate the 'somatic effective dose' the weighting factors recommended by ICRP 26 have been separated into those for somatic effects and for genetic effects. Calculation of the effective dose in diagnostic radiology procedures must consider the various technical parameters which determine the absorbed dose in the various organs, i.e. beam quality, typical entrance dose and the number of films of each view. Knowledge about these parameters is not always well established and therefore the effective dose estimates are very uncertain. The average dose absorbed by various organs in the case of administration of radionuclides to the body depends to a much higher degree on biological parameters than in the case of external irradiation. In contrast to the variability and lack of reliability of biological data, the physical methods for internal dose calculation are quite elaborate. However, these methods have to be extended to involve the target dose from the radioactivity distributed within the remaining parts of the body. An attempt was made to estimate the somatic effective dose for the most common diagnostic X-ray and nuclear medicine procedures. This would make it possible to compare the risk of X-ray and nuclear medicine techniques on a more equitable basis. The collective effective dose from medical irradiation is estimated for various countries on the basis of reported statistical data. (H.K.)

  4. A GPU-based Monte Carlo dose calculation code for photon transport in a voxel phantom

    International Nuclear Information System (INIS)

    Bellezzo, M.; Do Nascimento, E.; Yoriyaz, H.

    2014-08-01

    As the most accurate method to estimate absorbed dose in radiotherapy, Monte Carlo method has been widely used in radiotherapy treatment planning. Nevertheless, its efficiency can be improved for clinical routine applications. In this paper, we present the CUBMC code, a GPU-based Mc photon transport algorithm for dose calculation under the Compute Unified Device Architecture platform. The simulation of physical events is based on the algorithm used in Penelope, and the cross section table used is the one generated by the Material routine, als present in Penelope code. Photons are transported in voxel-based geometries with different compositions. To demonstrate the capabilities of the algorithm developed in the present work four 128 x 128 x 128 voxel phantoms have been considered. One of them is composed by a homogeneous water-based media, the second is composed by bone, the third is composed by lung and the fourth is composed by a heterogeneous bone and vacuum geometry. Simulations were done considering a 6 MeV monoenergetic photon point source. There are two distinct approaches that were used for transport simulation. The first of them forces the photon to stop at every voxel frontier, the second one is the Woodcock method, where the photon stop in the frontier will be considered depending on the material changing across the photon travel line. Dose calculations using these methods are compared for validation with Penelope and MCNP5 codes. Speed-up factors are compared using a NVidia GTX 560-Ti GPU card against a 2.27 GHz Intel Xeon CPU processor. (Author)

  5. A GPU-based Monte Carlo dose calculation code for photon transport in a voxel phantom

    Energy Technology Data Exchange (ETDEWEB)

    Bellezzo, M.; Do Nascimento, E.; Yoriyaz, H., E-mail: mbellezzo@gmail.br [Instituto de Pesquisas Energeticas e Nucleares / CNEN, Av. Lineu Prestes 2242, Cidade Universitaria, 05508-000 Sao Paulo (Brazil)

    2014-08-15

    As the most accurate method to estimate absorbed dose in radiotherapy, Monte Carlo method has been widely used in radiotherapy treatment planning. Nevertheless, its efficiency can be improved for clinical routine applications. In this paper, we present the CUBMC code, a GPU-based Mc photon transport algorithm for dose calculation under the Compute Unified Device Architecture platform. The simulation of physical events is based on the algorithm used in Penelope, and the cross section table used is the one generated by the Material routine, als present in Penelope code. Photons are transported in voxel-based geometries with different compositions. To demonstrate the capabilities of the algorithm developed in the present work four 128 x 128 x 128 voxel phantoms have been considered. One of them is composed by a homogeneous water-based media, the second is composed by bone, the third is composed by lung and the fourth is composed by a heterogeneous bone and vacuum geometry. Simulations were done considering a 6 MeV monoenergetic photon point source. There are two distinct approaches that were used for transport simulation. The first of them forces the photon to stop at every voxel frontier, the second one is the Woodcock method, where the photon stop in the frontier will be considered depending on the material changing across the photon travel line. Dose calculations using these methods are compared for validation with Penelope and MCNP5 codes. Speed-up factors are compared using a NVidia GTX 560-Ti GPU card against a 2.27 GHz Intel Xeon CPU processor. (Author)

  6. Stepped-irradiation SAR: A viable approach to circumvent OSL equivalent dose underestimation in last glacial loess of northwestern China

    International Nuclear Information System (INIS)

    Qin, J.T.; Zhou, L.P.

    2009-01-01

    The equivalent dose (D e ) obtained with the continuous irradiation SAR (CI-SAR) protocol for fine-grained quartz from loess of northwestern China is found to be lower than the expected value for samples older than 70 ka based on the regional stratigraphy. This is attributed to the difference in the response of the quartz to natural radiation and laboratory beta irradiation whose rates vary by ∼10 8 times. A stepped irradiation SAR protocol was employed to evaluate the influence of such a 'dose rate effect' on the equivalent dose determination. After investigating the effects of thermal treatment and 'unit-dose' on OSL signal and D e , we refined the stepped irradiation strategy with a 'unit-dose' of ∼25 Gy and successive thermal treatments at 250 deg. C for 10 s, and applied it to the SAR protocol. This stepped irradiation SAR (SI-SAR) protocol led to a 20%-70% increase in D e value for loess deposited during the early last glacial period.

  7. A point-kernel shielding code for calculations of neutron and secondary gamma-ray 1cm dose equivalents: PKN

    International Nuclear Information System (INIS)

    Kotegawa, Hiroshi; Tanaka, Shun-ichi

    1991-09-01

    A point-kernel integral technique code, PKN, and the related data library have been developed to calculate neutron and secondary gamma-ray dose equivalents in water, concrete and iron shields for neutron sources in 3-dimensional geometry. The comparison between calculational results of the present code and those of the 1-dimensional transport code ANISN = JR, and the 2-dimensional transport code DOT4.2 showed a sufficient accuracy, and the availability of the PKN code has been confirmed. (author)

  8. Ambient dose equivalent H*(d) - an appropriate philosophy for radiation monitoring onboard aircraft and in space?

    International Nuclear Information System (INIS)

    Vana, N.; Hajek, M.; Berger, T.

    2003-01-01

    In this paper authors deals with the ambient dose equivalent H * (d) and their application for onboard Aircraft and Space station. The discussion and the carried out experiments demonstrated that the philosophy of H * (10) leads to an underestimation of the whole-body radiation exposure when applied onboard aircraft and in space. It therefore has to be considered to introduce a new concept that could be based on microdosimetric principles, offering the unique potential of a more direct correlation to radiobiological parameters

  9. The paradox of human equivalent dose formula: A canonical case study of abrus precatorius aqueous leaf extract in monogastric animals

    Directory of Open Access Journals (Sweden)

    Saganuwan Alhaji Saganuwan

    2016-03-01

    Full Text Available There is abundant literature on the toxicity of A. precatorius seeds. However there is a need to define the toxicity limit of the Abrus precatorius leaf in monogastric animals. Human Equivalent Dose (HED which is equal to animal dose multiplied by animal km (metabolism constant divided by human km was used to project the LD50 of fifteen monogastric animals , where human km factor is body weight (kg divided by body surface area (m2. Human Equivalent No-observable Adverse Effect Doses were determined by multiplying the animal no-observable adverse effect dose by animal weight (Wa divided by human weight (Wh. The LD50 of the aqueous leaf extract of Abrus precatorius in mice was estimated to be between 2559.5 and 3123.3 mg/kg body weight. The LD50 extrapolated from mouse to rat (1349.3-1646.6 mg/kg, hamster (1855.3-2264.1 mg/kg, guinea pig (1279.5-1561.4 mg/kg, rabbit (618.4-754.7 mg/kg, monkey (593.7-724.5 mg/kg, cat (392.7-479.2 mg/kg, dog and baboon (371.1-452.8 mg/kg, child (297-362 mg/kg and adult human (197.8-241.5 mg/kg body weight respectively could be a reality. The therapeutic safe dose range for the animals was 1-12.5 mg/kg body weight for a period of 7 days, but at a dose (≤ 200 mg/kg body weight the leaf extract showed haematinic effect. However, at a higher dose (> 200 mg/kg, the extract showed haemolytic activity in rats, whereas at a dose (≥25.0 mg/kg, the leaf extract might be organotoxic in hamster, guinea pig, rabbit, monkey, cat, dog, baboon, child and adult human if administered orally for a period of 7 days.

  10. Evaluation of dose equivalent and identification of energy rangeby the electronic personal dosemeter for neutron 'Saphydose-N' at different workplaces of nuclear facilities

    International Nuclear Information System (INIS)

    Chau, Q.; Lahaye, T.; Lacoste, V.; Muller, H.; Luszik-Bhadra, M.; Reginatto, M.

    2005-01-01

    Full text: According to the transposition of the Directive 96/29/EURATOM into the French legislation, any worker operating in a controlled area has to be monitored by 'passive' and 'active' dosemeters. Electronic personal dosemeters are especially needed for optimization of workplaces. In nuclear facilities, some of the workers are likely to be exposed to mixed neutron-photon fields. If the dosimetry of the photons is relatively well controlled, the neutron dosimetry raises more difficulties. In this context, the Institute for Radiological Protection and Nuclear Safety (IRSN) has developed an electronic device based on a silicon detector: the 'Saphydose-N' dosemeter. It is composed of several detectors and covers more than eight energy decades. The Saphydose-N device complies with recommendations of standard IEC 1323. As the radiation fields produced in laboratory are not strictly representative to those encountered at workplaces, IRSN evaluated the response of Saphydose-N at various workplaces from nuclear industry. Several campaigns were performed in the framework of the European contract EVIDOS ('Evaluation of Individual Dosimetry in Mixed Neutron and Photon Radiation Fields'). Saphydose-N was carried out at a fuel processing plant, at the Nuclear Power Plant close to the boiling water reactor, near the transport and storage cask and at the Research Reactor. This paper mainly aims at presenting the measurements, the identification of the energy range and the results obtained with the electronic personal neutron Saphydose-N at different facilities. The results are compared with reference values, defined by other partners in EVIDOS project, for ambient and personal dose equivalent. (author)

  11. Calculation of equivalent dose index for electrons from 5,0 to 22,0 MeV by the Monte Carlo method

    International Nuclear Information System (INIS)

    Peixoto, J.E.

    1979-01-01

    The index of equivalent dose in depth and in a sphere surface of a soft tissue equivalent material were determined by Monte Carlo method for electron irradiations from 5,0 to 22.00 MeV. The effect of different irradiation geometries which simulate the incidence of onedirectional opposite rotational and isotropic beams was studied. It is also shown that the detector of wall thickness with 0.5g/cm 2 and isotropic response com be used to measure index of equivalent dose for fast electrons. The alternative concept of average equivalent dose for radiation protection is discussed. (M.C.K.) [pt

  12. Off-axis dose equivalent due to secondary neutrons from uniform scanning proton beams during proton radiotherapy

    Science.gov (United States)

    Islam, M. R.; Collums, T. L.; Zheng, Y.; Monson, J.; Benton, E. R.

    2013-11-01

    The production of secondary neutrons is an undesirable byproduct of proton therapy and it is important to quantify the contribution from secondary neutrons to patient dose received outside the treatment volume. The purpose of this study is to investigate the off-axis dose equivalent from secondary neutrons experimentally using CR-39 plastic nuclear track detectors (PNTD) at ProCure Proton Therapy Center, Oklahoma City, OK. In this experiment, we placed several layers of CR-39 PNTD laterally outside the treatment volume inside a phantom and in air at various depths and angles with respect to the primary beam axis. Three different proton beams with max energies of 78, 162 and 226 MeV and 4 cm modulation width, a 5 cm diameter brass aperture, and a small snout located 38 cm from isocenter were used for the entire experiment. Monte Carlo simulations were also performed based on the experimental setup using a simplified snout configuration and the FLUKA Monte Carlo radiation transport code. The measured ratio of secondary neutron dose equivalent to therapeutic primary proton dose (H/D) ranged from 0.3 ± 0.08 mSv Gy-1 for 78 MeV proton beam to 37.4 ± 2.42 mSv Gy-1 for 226 MeV proton beam. Both experiment and simulation showed a similar decreasing trend in dose equivalent with distance to the central axis and the magnitude varied by a factor of about 2 in most locations. H/D was found to increase as the energy of the primary proton beam increased and higher H/D was observed at 135° compared to 45° and 90°. The overall higher H/D in air indicates the predominance of external neutrons produced in the nozzle rather than inside the body.

  13. Off-axis dose equivalent due to secondary neutrons from uniform scanning proton beams during proton radiotherapy

    International Nuclear Information System (INIS)

    Islam, M R; Collums, T L; Monson, J; Benton, E R; Zheng, Y

    2013-01-01

    The production of secondary neutrons is an undesirable byproduct of proton therapy and it is important to quantify the contribution from secondary neutrons to patient dose received outside the treatment volume. The purpose of this study is to investigate the off-axis dose equivalent from secondary neutrons experimentally using CR-39 plastic nuclear track detectors (PNTD) at ProCure Proton Therapy Center, Oklahoma City, OK. In this experiment, we placed several layers of CR-39 PNTD laterally outside the treatment volume inside a phantom and in air at various depths and angles with respect to the primary beam axis. Three different proton beams with max energies of 78, 162 and 226 MeV and 4 cm modulation width, a 5 cm diameter brass aperture, and a small snout located 38 cm from isocenter were used for the entire experiment. Monte Carlo simulations were also performed based on the experimental setup using a simplified snout configuration and the FLUKA Monte Carlo radiation transport code. The measured ratio of secondary neutron dose equivalent to therapeutic primary proton dose (H/D) ranged from 0.3 ± 0.08 mSv Gy −1  for 78 MeV proton beam to 37.4 ± 2.42 mSv Gy −1  for 226 MeV proton beam. Both experiment and simulation showed a similar decreasing trend in dose equivalent with distance to the central axis and the magnitude varied by a factor of about 2 in most locations. H/D was found to increase as the energy of the primary proton beam increased and higher H/D was observed at 135° compared to 45° and 90°. The overall higher H/D in air indicates the predominance of external neutrons produced in the nozzle rather than inside the body. (paper)

  14. A single-source photon source model of a linear accelerator for Monte Carlo dose calculation.

    Science.gov (United States)

    Nwankwo, Obioma; Glatting, Gerhard; Wenz, Frederik; Fleckenstein, Jens

    2017-01-01

    To introduce a new method of deriving a virtual source model (VSM) of a linear accelerator photon beam from a phase space file (PSF) for Monte Carlo (MC) dose calculation. A PSF of a 6 MV photon beam was generated by simulating the interactions of primary electrons with the relevant geometries of a Synergy linear accelerator (Elekta AB, Stockholm, Sweden) and recording the particles that reach a plane 16 cm downstream the electron source. Probability distribution functions (PDFs) for particle positions and energies were derived from the analysis of the PSF. These PDFs were implemented in the VSM using inverse transform sampling. To model particle directions, the phase space plane was divided into a regular square grid. Each element of the grid corresponds to an area of 1 mm2 in the phase space plane. The average direction cosines, Pearson correlation coefficient (PCC) between photon energies and their direction cosines, as well as the PCC between the direction cosines were calculated for each grid element. Weighted polynomial surfaces were then fitted to these 2D data. The weights are used to correct for heteroscedasticity across the phase space bins. The directions of the particles created by the VSM were calculated from these fitted functions. The VSM was validated against the PSF by comparing the doses calculated by the two methods for different square field sizes. The comparisons were performed with profile and gamma analyses. The doses calculated with the PSF and VSM agree to within 3% /1 mm (>95% pixel pass rate) for the evaluated fields. A new method of deriving a virtual photon source model of a linear accelerator from a PSF file for MC dose calculation was developed. Validation results show that the doses calculated with the VSM and the PSF agree to within 3% /1 mm.

  15. A single-source photon source model of a linear accelerator for Monte Carlo dose calculation.

    Directory of Open Access Journals (Sweden)

    Obioma Nwankwo

    Full Text Available To introduce a new method of deriving a virtual source model (VSM of a linear accelerator photon beam from a phase space file (PSF for Monte Carlo (MC dose calculation.A PSF of a 6 MV photon beam was generated by simulating the interactions of primary electrons with the relevant geometries of a Synergy linear accelerator (Elekta AB, Stockholm, Sweden and recording the particles that reach a plane 16 cm downstream the electron source. Probability distribution functions (PDFs for particle positions and energies were derived from the analysis of the PSF. These PDFs were implemented in the VSM using inverse transform sampling. To model particle directions, the phase space plane was divided into a regular square grid. Each element of the grid corresponds to an area of 1 mm2 in the phase space plane. The average direction cosines, Pearson correlation coefficient (PCC between photon energies and their direction cosines, as well as the PCC between the direction cosines were calculated for each grid element. Weighted polynomial surfaces were then fitted to these 2D data. The weights are used to correct for heteroscedasticity across the phase space bins. The directions of the particles created by the VSM were calculated from these fitted functions. The VSM was validated against the PSF by comparing the doses calculated by the two methods for different square field sizes. The comparisons were performed with profile and gamma analyses.The doses calculated with the PSF and VSM agree to within 3% /1 mm (>95% pixel pass rate for the evaluated fields.A new method of deriving a virtual photon source model of a linear accelerator from a PSF file for MC dose calculation was developed. Validation results show that the doses calculated with the VSM and the PSF agree to within 3% /1 mm.

  16. The intercomparison of 137Cs irradiator output measurement and personal dose equivalent, Hp(10), using TLD and film

    International Nuclear Information System (INIS)

    Nazaroh; Susetyo Trijoko; Sri Inang Sunaryati

    2010-01-01

    Intercomparison of output measurement of 137 Cs irradiator and personal dose equivalent, Hp(10) using TLD and film have been carried out in the year of 2006 to 2008. According to IAEA recommendation, intercomparison is one of audit activities but it is performed in the spirit of collaboration and support rather than in the spirit of inspection. The aim of intercomparison of output measurement of 137 Cs irradiator is to verify the dose stated by the participant laboratories. Intercomparison is also to assess the competency of the participant, to keep traceability and consistency of measurement result, to assure that instrument work correctly and the result of evaluation was in agreement, and also for fulfilling one of the clauses of ISO-17025-2005. Besides that, this intercomparison aimed to facilitate link between the system and members of national measurement and transfer of experience in measurement technique and dose evaluation of radiation. The benefit of intercomparison is important among others as tests of proficiency in dose evaluation or measurement, upgrading quality of service and for obeying supervisor body legislation (BAPETEN). TLD was used as a means of output 137 Cs irradiator measurement, whereas film and TLD were used for dose intercomparison. This paper presented result of intercomparison of output measurement and evaluation of personal dose equivalent, Hp(10) in the year of 2006 to 2008 followed by 6 participants: Balai Pengamanan Fasilitas Kesehatan (BPFK) Jakarta, Medan, Surabaya, Makasar, PTLR and Laboratory of Keselamatan dan Kesehatan Lingkungan (KKL)-PTKMR BATAN. In this intercomparison, the dose of TLD stated by participant were compared with the dose measured by Radiation Metrology Laboratory (LMR), and the results showed the differences were within 10 %, so it was satisfied. The results of intercomparison of personal dose equivalent, Hp(10) were evaluated based on ISO/IEC Guide 43-1, 1997 analysis and expressed as E n . The values of E n

  17. Calculation of conversion factor of Kerma in the air for ambient dose equivalent in radiotherapy

    International Nuclear Information System (INIS)

    Lima, Marco Antonio Frota

    2000-03-01

    This work aims to estimate the average conversion factor of Kerma in air to H * (10) using photon beams coming from clinic linear accelerators, transmitted through concrete walls of a radiotherapic treatment room. The transmitted photon spectra by both 1 meter and 2 meters concrete walls, in an area of 40 x 40 cm 2 , were calculated when the primary beam impart in an angle of 0 deg. The (secondary) photon beams transmitted respectively by 0,5 meter, 1,0 meter, 1,0 meter and 2,0 meter concrete walls, after they scattered by an angle of 90 deg in a cylindric phantom inside the room, were also determined. Generally, 50 millions of histories were computed for each simulation made for the primary beam. For the 90 deg spread, the number of histories was 100 millions. The computational code used on this work was the MCNP4B. The most common clinic accelerators used on radiotheraphic treatments were used on this work CLINAC-4, CLINAC-6, CLINAC-18 and CLINAC-2500. From the spectra analysis obtained in this work, it was possible to dispose the conversion factor for realistic beams found in radiotherapeutic establishment. (author)

  18. Study of the equivalent dose distribution in organs and tissues using periapical odontological radiography

    International Nuclear Information System (INIS)

    Santos, H.F.S.; Cipeli, J.F.; Fortes, M.A.B.; Federico, C.A.

    2017-01-01

    In this article presents a study of the doses obtained in periapical odontological radiography in main tissues of the head, using thermoluminescent dosemeters of type TLD-700H applied to a anthropomorphic simulator. The results indicate that the skin and salivary glands received the highest doses and the risk of calculated injury was 1.44 x 10 -6 Sv -1 per radiograph

  19. Different concentrations and volumes of p-phenylenediamine in pet. (equivalent doses) are associated with similar patch test outcomes

    DEFF Research Database (Denmark)

    Andersen, Flemming; Hamann, Carsten R; Andersen, Klaus E

    2018-01-01

    BACKGROUND: Concern about causing active sensitization when patch testing is performed with p-phenylenediamine (PPD) 1% pet. has led to a recommendation to use PPD 0.3% pet. as a potentially safer preparation. However, the dose per area of allergen delivered, and hence the risk of active...... sensitization, depend on the amount dispensed into the patch test chamber, which can vary widely. OBJECTIVE: To evaluate whether patch testing with equivalent doses of different concentrations of PPD in pet. is associated with similar outcomes. METHODS: Seventeen known PPD-sensitive subjects were patch tested...... with different volumes and concentrations of PPD in pet. that deliver the same allergen dose per unit area (6 mg of PPD 1% pet. and 20 mg of PPD 0.3% pet. in Finn Chambers®, both equivalent to ∼ 0.09 mg/cm2 ). RESULTS: Eleven patients (65%) had positive reactions to both doses; 4 patients (24%) had negative...

  20. Dose equivalent near the bone-soft tissue interface from nuclear fragments produced by high-energy protons

    Science.gov (United States)

    Shavers, M. R.; Poston, J. W.; Cucinotta, F. A.; Wilson, J. W.

    1996-01-01

    During manned space missions, high-energy nucleons of cosmic and solar origin collide with atomic nuclei of the human body and produce a broad linear energy transfer spectrum of secondary particles, called target fragments. These nuclear fragments are often more biologically harmful than the direct ionization of the incident nucleon. That these secondary particles increase tissue absorbed dose in regions adjacent to the bone-soft tissue interface was demonstrated in a previous publication. To assess radiological risks to tissue near the bone-soft tissue interface, a computer transport model for nuclear fragments produced by high energy nucleons was used in this study to calculate integral linear energy transfer spectra and dose equivalents resulting from nuclear collisions of 1-GeV protons transversing bone and red bone marrow. In terms of dose equivalent averaged over trabecular bone marrow, target fragments emitted from interactions in both tissues are predicted to be at least as important as the direct ionization of the primary protons-twice as important, if recently recommended radiation weighting factors and "worst-case" geometry are used. The use of conventional dosimetry (absorbed dose weighted by aa linear energy transfer-dependent quality factor) as an appropriate framework for predicting risk from low fluences of high-linear energy transfer target fragments is discussed.

  1. Evaluation of the effective equivalent dose in the general public due to the discharge of uranium in groundwater

    International Nuclear Information System (INIS)

    Gordon, A.M.P.L.; Jacomino, V.M.F.

    1989-12-01

    Some facilities available at IPEN-CNEN/SP may discharge uranium in their liquid effluents. The uranium contents of these effluents are analyzed by photometry or fluorimetry, and according to the results obtained a decision is made, by the Environmental Monitoring Division, upon their discharge to the environment. In 1988 a total activity of 3.66x10 9 Bq of uranium was discharge in a volume of approximately 30 m 3 . The effective equivalent dose in the general public was evaluated by making a conservative assumption that all the liquid effluents containing uranium are discharged directly to the soil reaching the groundwater. The dose calculation was carried out by using a generic model which described the transport of radionuclides in the groundwater. In order to be conservative it was also assumed that the critical pathway is the direct in gestion of water through hypothetical wells around the Institute. Conservative assumptions were also made in the characterization of the local aquifer parameters such as vertical and longitudinal dispersivity, effective porosity of the soil, hydraulic conductivity etc., in roder to overestimate the effective equivalent dose. The result obtained, of 5.3x10 -10 mSv/a is far below the dose limit for the public adopted by the Radiological Protection Board. The derived limit for the discharge was also evaluated, using the same model, giving a result of 3.6x10 13 Bq/a. (author) [pt

  2. Photonics

    CERN Document Server

    Andrews, David L

    2015-01-01

    Discusses the basic physical principles underlying Biomedical Photonics, spectroscopy and microscopy This volume discusses biomedical photonics, spectroscopy and microscopy, the basic physical principles underlying the technology and its applications. The topics discussed in this volume are: Biophotonics; Fluorescence and Phosphorescence; Medical Photonics; Microscopy; Nonlinear Optics; Ophthalmic Technology; Optical Tomography; Optofluidics; Photodynamic Therapy; Image Processing; Imaging Systems; Sensors; Single Molecule Detection; Futurology in Photonics. Comprehensive and accessible cov

  3. Photonics

    CERN Document Server

    Andrews, David L

    2015-01-01

    Discusses the basic physical principles underlying the technology instrumentation of photonics This volume discusses photonics technology and instrumentation. The topics discussed in this volume are: Communication Networks; Data Buffers; Defense and Security Applications; Detectors; Fiber Optics and Amplifiers; Green Photonics; Instrumentation and Metrology; Interferometers; Light-Harvesting Materials; Logic Devices; Optical Communications; Remote Sensing; Solar Energy; Solid-State Lighting; Wavelength Conversion Comprehensive and accessible coverage of the whole of modern photonics Emphas

  4. Photonics

    CERN Document Server

    Andrews, David L

    2015-01-01

    Discusses the basic physical principles underlying thescience and technology of nanophotonics, its materials andstructures This volume presents nanophotonic structures and Materials.Nanophotonics is photonic science and technology that utilizeslight/matter interactions on the nanoscale where researchers arediscovering new phenomena and developing techniques that go wellbeyond what is possible with conventional photonics andelectronics.The topics discussed in this volume are: CavityPhotonics; Cold Atoms and Bose-Einstein Condensates; Displays;E-paper; Graphene; Integrated Photonics; Liquid Cry

  5. Photonics

    CERN Document Server

    Andrews, David L

    2015-01-01

    This book covers modern photonics accessibly and discusses the basic physical principles underlying all the applications and technology of photonicsThis volume covers the basic physical principles underlying the technology and all applications of photonics from statistical optics to quantum optics. The topics discussed in this volume are: Photons in perspective; Coherence and Statistical Optics; Complex Light and Singular Optics; Electrodynamics of Dielectric Media; Fast and slow Light; Holography; Multiphoton Processes; Optical Angular Momentum; Optical Forces, Trapping and Manipulation; Pol

  6. Emphysema quantification and lung volumetry in chest X-ray equivalent ultralow dose CT - Intra-individual comparison with standard dose CT.

    Science.gov (United States)

    Messerli, Michael; Ottilinger, Thorsten; Warschkow, René; Leschka, Sebastian; Alkadhi, Hatem; Wildermuth, Simon; Bauer, Ralf W

    2017-06-01

    To determine whether ultralow dose chest CT with tin filtration can be used for emphysema quantification and lung volumetry and to assess differences in emphysema measurements and lung volume between standard dose and ultralow dose CT scans using advanced modeled iterative reconstruction (ADMIRE). 84 consecutive patients from a prospective, IRB-approved single-center study were included and underwent clinically indicated standard dose chest CT (1.7±0.6mSv) and additional single-energy ultralow dose CT (0.14±0.01mSv) at 100kV and fixed tube current at 70mAs with tin filtration in the same session. Forty of the 84 patients (48%) had no emphysema, 44 (52%) had emphysema. One radiologist performed fully automated software-based pulmonary emphysema quantification and lung volumetry of standard and ultralow dose CT with different levels of ADMIRE. Friedman test and Wilcoxon rank sum test were used for multiple comparison of emphysema and lung volume. Lung volumes were compared using the concordance correlation coefficient. The median low-attenuation areas (LAA) using filtered back projection (FBP) in standard dose was 4.4% and decreased to 2.6%, 2.1% and 1.8% using ADMIRE 3, 4, and 5, respectively. The median values of LAA in ultralow dose CT were 5.7%, 4.1% and 2.4% for ADMIRE 3, 4, and 5, respectively. There was no statistically significant difference between LAA in standard dose CT using FBP and ultralow dose using ADMIRE 4 (p=0.358) as well as in standard dose CT using ADMIRE 3 and ultralow dose using ADMIRE 5 (p=0.966). In comparison with standard dose FBP the concordance correlation coefficients of lung volumetry were 1.000, 0.999, and 0.999 for ADMIRE 3, 4, and 5 in standard dose, and 0.972 for ADMIRE 3, 4 and 5 in ultralow dose CT. Ultralow dose CT at chest X-ray equivalent dose levels allows for lung volumetry as well as detection and quantification of emphysema. However, longitudinal emphysema analyses should be performed with the same scan protocol and

  7. On the absorbed dose determination method in high energy photon beams

    International Nuclear Information System (INIS)

    Scarlat, F.; Scarisoreanu, A.; Oane, M.; Mitru, E.; Avadanei, C.

    2008-01-01

    The absorbed dose determination method in water, based on standards of air kerma or exposure in high energy photon beams generated by electron with energies in the range of 1 MeV to 50 MeV is presented herein. The method is based on IAEA-398, AAPM TG-51, DIN 6800-2, IAEA-381, IAEA-277 and NACP-80 recommendations. The dosimetry equipment is composed of UNIDOS T 10005 electrometer and different ionization chambers calibrated in air kerma method in a Co 60 beam. Starting from the general formalism showed in IAEA-381, the determination of absorbed dose in water, under reference conditions in high energy photon beams, is given. This method was adopted for the secondary standard dosimetry laboratory (SSDL) in NILPRP-Bucharest

  8. Survival of tumor bearing mice by sequencing of low dose rate (LDR) neutron and photon radiation

    International Nuclear Information System (INIS)

    Onomura, C.I.; Feola, J.M.; Maruyama, Y.

    1984-01-01

    Cf-252 neutron radiation (NT) has been shown to be effective therapy for bulky, hypoxic human tumor and to produce consistent rapid clearance and 5 year cures. NT has been found to be more or less effective depending upon the schedule in which it is used and upon mixing with photon radiation. In an effort to study this scheduling and photon effect, LSA tumor was irradiated in vivo in a hypoxic, advanced state, in different schedules in combination of NT with Co-60 photons. The LSA lymphoma of C57BL/ym mice represents an accurate system to assess dose-response of tumor cells in vivo. Mean survival time was used as endpoint. A high RBE for LDR Cf-252 NT was observed with a RBE(n) of -- 5.0. The effect was not greatly sensitive to sequence in which photons were used. Comparison studies were also tested relative to LDR Cs-137 photon radiation. The results support the high efficacy of LDR NT for destruction of hypoxic tumor in vivo

  9. Estimation of collective effective dose equivalent from environmental radiation and radioactive materials in Japan. A preliminary study

    International Nuclear Information System (INIS)

    Maruyama, Takashi; Noda, Yutaka; Takeshita, Mitsue; Iwai, Kazuo.

    1994-01-01

    The peaceful uses of nuclear power and radiations have been developed into a stage of practical applications for human life. Radiation causes harmful effects to human beings, although human beings receives a number of invaluable benefits from the nuclear energy and the uses of radiation. In order to examine the optimization of radiation protection in these practices, collective effective dose equivalent from environmental exposures due to natural and artificial radiations have been preliminarily evaluated using most recent data. The resultant collective doses were compared with those from medical and occupational exposures. It is noted that, in Japan, the collective effective dose from environmental radiation sources can be approximately same to that from medical exposure. (author)

  10. Dose response of commercially available optically stimulated luminescent detector, Al2O3:C for megavoltage photons and electrons.

    Science.gov (United States)

    Kim, Dong Wook; Chung, Weon Kuu; Shin, Dong Oh; Yoon, Myonggeun; Hwang, Ui-Jung; Rah, Jeong-Eun; Jeong, Hojin; Lee, Sang Yeob; Shin, Dongho; Lee, Se Byeong; Park, Sung Yong

    2012-04-01

    This study examined the dose response of an optically stimulated luminescence dosemeter (OSLD) to megavoltage photon and electron beams. A nanoDot™ dosemeter was used to measure the dose response of the OSLD. Photons of 6-15 MV and electrons of 9-20 MeV were delivered by a Varian 21iX machine (Varian Medical System, Inc. Milpitas, CA, USA). The energy dependency was dose was linear until 200 cGy. The superficial dose measurements revealed photon irradiation to have an angular dependency. The nanoDot™ dosemeter has potential use as an in vivo dosimetric tool that is independent of the energy, has dose linearity and a rapid response compared with normal in vivo dosimetric tools, such as thermoluminescence detectors. However, the OSLD must be treated very carefully due to the high angular dependency of the photon beam.

  11. Photon activation therapy: a Monte Carlo study on dose enhancement by various sources and activation media

    International Nuclear Information System (INIS)

    Bakhshabadi, Mahdi; Ghorbani, Mahdi; Meigooni, Ali Soleimani

    2013-01-01

    In the present study, a number of brachytherapy sources and activation media were simulated using MCNPX code and the results were analyzed based on the dose enhancement factor values. Furthermore, two new brachytherapy sources ( 131 Cs and a hypothetical 170 Tm) were evaluated for their application in photon activation therapy (PAT). 125 I, 103 Pd, 131 Cs and hypothetical 170 Tm brachytherapy sources were simulated in water and their dose rate constant and the radial dose functions were compared with previously published data. The sources were then simulated in a soft tissue phantom which was composed of Ag, I, Pt or Au as activation media uniformly distributed in the tumour volume. These simulations were performed using the MCNPX code, and dose enhancement factor (DEF) was obtained for 7, 18 and 30 mg/ml concentrations of the activation media. Each source, activation medium and concentration was evaluated separately in a separate simulation. The calculated dose rate constant and radial dose functions were in agreement with the published data for the aforementioned sources. The maximum DEF was found to be 5.58 for a combination of the 170 Tm source with 30 mg/ml concentration of I. The DEFs for 131 Cs and 170 Tm sources for all the four activation media were higher than those for other sources and activation media. From this point of view, these two sources can be more useful in photon activation therapy with photon emitter sources. Furthermore, 131 Cs and 170 Tm brachytherapy sources can be proposed as new options for use in the field of PAT.

  12. Projected Second Tumor Risk and Dose to Neurocognitive Structures After Proton Versus Photon Radiotherapy for Benign Meningioma

    Energy Technology Data Exchange (ETDEWEB)

    Arvold, Nils D. [Harvard Radiation Oncology Program, Harvard Medical School, Boston, MA (United States); Niemierko, Andrzej; Broussard, George P.; Adams, Judith; Fullerton, Barbara; Loeffler, Jay S. [Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA (United States); Shih, Helen A., E-mail: hshih@partners.org [Department of Radiation Oncology, Massachusetts General Hospital and Harvard Medical School, Boston, MA (United States)

    2012-07-15

    Purpose: To calculated projected second tumor rates and dose to organs at risk (OAR) in patients with benign intracranial meningioma (BM), according to dosimetric comparisons between proton radiotherapy (PRT) and photon radiotherapy (XRT) treatment plans. Methods and Materials: Ten patients with BM treated at Massachusetts General Hospital during 2006-2010 with PRT were replanned with XRT (intensity-modulated or three-dimensional conformal radiotherapy), optimizing dose to the tumor while sparing OAR. Total dose was 54 Gy in 1.8 Gy per fraction for all plans. We calculated equivalent uniform doses, normal tissue complication probabilities, and whole brain-based estimates of excess risk of radiation-associated intracranial second tumors. Results: Excess risk of second tumors was significantly lower among PRT compared with XRT plans (1.3 vs. 2.8 per 10,000 patients per year, p < 0.002). Mean equivalent uniform doses were lower among PRT plans for the whole brain (19.0 vs. 22.8 Gy, p < 0.0001), brainstem (23.8 vs. 35.2 Gy, p = 0.004), hippocampi (left, 13.5 vs. 25.6 Gy, p < 0.0001; right, 7.6 vs. 21.8 Gy, p = 0.001), temporal lobes (left, 25.8 vs. 34.6 Gy, p = 0.007; right, 25.8 vs. 32.9 Gy, p = 0.008), pituitary gland (29.2 vs. 37.0 Gy, p = 0.047), optic nerves (left, 28.5 vs. 33.8 Gy, p = 0.04; right, 25.1 vs. 31.1 Gy, p = 0.07), and cochleas (left, 12.2 vs. 15.8 Gy, p = 0.39; right,1.5 vs. 8.8 Gy, p = 0.01). Mean normal tissue complication probability was <1% for all structures and not significantly different between PRT and XRT plans. Conclusions: Compared with XRT, PRT for BM decreases the risk of RT-associated second tumors by half and delivers significantly lower doses to neurocognitive and critical structures of vision and hearing.

  13. Projected Second Tumor Risk and Dose to Neurocognitive Structures After Proton Versus Photon Radiotherapy for Benign Meningioma

    International Nuclear Information System (INIS)

    Arvold, Nils D.; Niemierko, Andrzej; Broussard, George P.; Adams, Judith; Fullerton, Barbara; Loeffler, Jay S.; Shih, Helen A.

    2012-01-01

    Purpose: To calculated projected second tumor rates and dose to organs at risk (OAR) in patients with benign intracranial meningioma (BM), according to dosimetric comparisons between proton radiotherapy (PRT) and photon radiotherapy (XRT) treatment plans. Methods and Materials: Ten patients with BM treated at Massachusetts General Hospital during 2006–2010 with PRT were replanned with XRT (intensity-modulated or three-dimensional conformal radiotherapy), optimizing dose to the tumor while sparing OAR. Total dose was 54 Gy in 1.8 Gy per fraction for all plans. We calculated equivalent uniform doses, normal tissue complication probabilities, and whole brain–based estimates of excess risk of radiation-associated intracranial second tumors. Results: Excess risk of second tumors was significantly lower among PRT compared with XRT plans (1.3 vs. 2.8 per 10,000 patients per year, p < 0.002). Mean equivalent uniform doses were lower among PRT plans for the whole brain (19.0 vs. 22.8 Gy, p < 0.0001), brainstem (23.8 vs. 35.2 Gy, p = 0.004), hippocampi (left, 13.5 vs. 25.6 Gy, p < 0.0001; right, 7.6 vs. 21.8 Gy, p = 0.001), temporal lobes (left, 25.8 vs. 34.6 Gy, p = 0.007; right, 25.8 vs. 32.9 Gy, p = 0.008), pituitary gland (29.2 vs. 37.0 Gy, p = 0.047), optic nerves (left, 28.5 vs. 33.8 Gy, p = 0.04; right, 25.1 vs. 31.1 Gy, p = 0.07), and cochleas (left, 12.2 vs. 15.8 Gy, p = 0.39; right,1.5 vs. 8.8 Gy, p = 0.01). Mean normal tissue complication probability was <1% for all structures and not significantly different between PRT and XRT plans. Conclusions: Compared with XRT, PRT for BM decreases the risk of RT-associated second tumors by half and delivers significantly lower doses to neurocognitive and critical structures of vision and hearing.

  14. An overview of equivalent doses in eye lens of occupational radiation workers in medical, industrial and nuclear areas

    International Nuclear Information System (INIS)

    Lima, A.R.; Silva, F.C.A. da; Hunt, J.G.

    2013-01-01

    Some epidemiological evidences were recently reviewed by the ICRP and it was suggested that, for the eye lens, the absorbed dose threshold for induction of late detriments is about 0.5 Gy. On this basis, on 2011, the ICRP has recommended changes to the occupational dose limit in planned exposure situations, reducing the eye lens dose equivalent limit of 150 mSv to 20 mSv per year, on average, during the period of 5 years, with exposure not exceeding 50 mSv in a single year. Following the ICRP recommendation, the Brazilian Commission of Nuclear Energy (CNEN) adopted immediately the new limit to the eyes lens. This study aimed to show an overview about the doses in eye lens of occupational radiation workers in situations of planned exposures in the medical, industrial and nuclear areas, emphasizing the greatest radiological risks applications. It was observed that there are some limitations, such as example, to use individual monitor calibrated on Hp(3), to assess the equivalent dose in the eye lens. This limitation obstructs some experimental studies and monitoring of the levels of radiation received in the eye lens of radiation workers. Recent studies have showed that the lenses of eyes monitoring of workers, mainly in the planned exposure, must be follow-up. However, such researches were obtained only in medical exposures, mainly in interventional medicine procedures. Studies with planned exposure on nuclear and industrial areas are really needed and will be very important due to the new recommended by ICRP dose limits. (author)

  15. A photon spectrometric dose-rate constant determination for the Advantage™ Pd-103 brachytherapy source

    OpenAIRE

    Chen, Zhe Jay; Bongiorni, Paul; Nath, Ravinder

    2010-01-01

    Purpose: Although several dosimetric characterizations using Monte Carlo simulation and thermoluminescent dosimetry (TLD) have been reported for the new Advantage™ Pd-103 source (IsoAid, LLC, Port Richey, FL), no AAPM consensus value has been established for the dosimetric parameters of the source. The aim of this work was to perform an additional dose-rate constant (Λ) determination using a recently established photon spectrometry technique (PST) that is independent of the published TLD and ...

  16. Dose rate considerations in brachytherapy: biological equivalence of LDR and HDR

    International Nuclear Information System (INIS)

    Orton, C.G.

    1994-01-01

    The linear-quadratic model for cell survival and bioeffect doses is discussed and equations for low dose rate (LDR), high dose rate (HDR) and intermediate situations are presented. The model, when used to define LDR and single fractions of HDR, shows, that these correspond to irradiations lasting longer than about 14 hours or shorter than about 0.7 hours, respectively. It is shown that, for HDR to be as safe and effective as LDR, the dose-rate effect of LDR has to be replaced by the fractionation-effect of HDR. This is necessary in order to take advantage of the differential repair characteristics between late-reacting normal tissue and tumor cells at low doses and low dose rates. Using the linear-quadratic model to simulate repair mathematically, it is shown that the number of fractions required is highly dependent upon what parameters are assumed for normal tissues and tumor, as well as whether or not there is any physical advantage gained by conversion from LDR to HDR. (author). 20 refs., 7 figs

  17. Evaluation of ambient dose equivalent rates influenced by vertical and horizontal distribution of radioactive cesium in soil in Fukushima Prefecture.

    Science.gov (United States)

    Malins, Alex; Kurikami, Hiroshi; Nakama, Shigeo; Saito, Tatsuo; Okumura, Masahiko; Machida, Masahiko; Kitamura, Akihiro

    2016-01-01

    The air dose rate in an environment contaminated with (134)Cs and (137)Cs depends on the amount, depth profile and horizontal distribution of these contaminants within the ground. This paper introduces and verifies a tool that models these variables and calculates ambient dose equivalent rates at 1 m above the ground. Good correlation is found between predicted dose rates and dose rates measured with survey meters in Fukushima Prefecture in areas contaminated with radiocesium from the Fukushima Dai-ichi Nuclear Power Plant accident. This finding is insensitive to the choice for modeling the activity depth distribution in the ground using activity measurements of collected soil layers, or by using exponential and hyperbolic secant fits to the measurement data. Better predictions are obtained by modeling the horizontal distribution of radioactive cesium across an area if multiple soil samples are available, as opposed to assuming a spatially homogeneous contamination distribution. Reductions seen in air dose rates above flat, undisturbed fields in Fukushima Prefecture are consistent with decrement by radioactive decay and downward migration of cesium into soil. Analysis of remediation strategies for farmland soils confirmed that topsoil removal and interchanging a topsoil layer with a subsoil layer result in similar reductions in the air dose rate. These two strategies are more effective than reverse tillage to invert and mix the topsoil. Copyright © 2015 Elsevier Ltd. All rights reserved.

  18. Application of biological effective dose (BED) to estimate the duration of symptomatic relief and repopulation dose equivalent in palliative radiotherapy and chemotherapy

    International Nuclear Information System (INIS)

    Jones, Bleddyn; Cominos, Matilda; Dale, Roger G.

    2003-01-01

    Purpose: To investigate the potential for mathematic modeling in the assessment of symptom relief in palliative radiotherapy and cytotoxic chemotherapy. Methods: The linear quadratic model of radiation effect with the overall treatment time and the daily dose equivalent of repopulation is modified to include the regrowth time after completion of therapy. Results: The predicted times to restore the original tumor volumes after treatment are dependent on the biological effective dose (BED) delivered and the repopulation parameter (K); it is also possible to estimate K values from analysis of palliative treatment response durations. Hypofractionated radiotherapy given at a low total dose may produce long symptom relief in slow-growing tumors because of their low α/β ratios (which confer high fraction sensitivity) and their slow regrowth rates. Cancers that have high α/β ratios (which confer low fraction sensitivity), and that are expected to repopulate rapidly during therapy, are predicted to have short durations of symptom control. The BED concept can be used to estimate the equivalent dose of radiotherapy that will achieve the same duration of symptom relief as palliative chemotherapy. Conclusion: Relatively simple radiobiologic modeling can be used to guide decision-making regarding the choice of the most appropriate palliative schedules and has important implications in the design of radiotherapy or chemotherapy clinical trials. The methods described provide a rationalization for treatment selection in a wide variety of tumors

  19. Derivation, evidence and physical validity of a weighted beam-zone method for dose determination in blocked photon fields of arbitrary shape

    International Nuclear Information System (INIS)

    Glaeser, L.; Quast, U.

    1981-01-01

    A simple, practical procedure for dose determination at any point of an arbitrarily shaped field has been derived: Square-field photon beams are sectioned into a set of pyramid-shell-like parts (beam zones), nested into each other around the smallest realizable square field, of different sizes but with equal dose contributions (thus weighted) with respect to a central dose reference point. The dose at any reference point in an irregular field can be determined simply by counting the number of non-shielded dose-contributing zones (or zone fractions), leading to the associated order of square-field size (with the same number of zones), the equivalent field with known dose. For experimental evidence of the validity of the weighted beam-zone method, measurements were carried out with different high-energy photon beams with one or more beam zones shielded by absorbing blocks. Measurements were made at points in unshielded and shielded parts of the field, on and off the beam axis and at different depths in a phantom. Calculations and measurements were compared. While relative depth doses were shown to be equal to within +-2% over a range of 5 cm ahead of and behind the dose reference point, the absolute dose deviations were within +-4%. The sources of error were analysed. They were mainly determined by scattered radiation from the beam limiting device and the partial shielding deriving from the shielding blocks. The same errors also occur in most of the known methods of dose calculation in irregular fields. (author)

  20. Nuclear mass inventory, photon dose rate and thermal decay heat of spent research reactor fuel assemblies

    International Nuclear Information System (INIS)

    Pond, R.B.; Matos, J.E.

    1996-05-01

    As part of the Department of Energy's spent nuclear fuel acceptance criteria, the mass of uranium and transuranic elements in spent research reactor fuel must be specified. These data are, however, not always known or readily determined. It is the purpose of this report to provide estimates of these data for some of the more common research reactor fuel assembly types. The specific types considered here are MTR, TRIGA and DIDO fuel assemblies. The degree of physical protection given to spent fuel assemblies is largely dependent upon the photon dose rate of the spent fuel material. These data also, are not always known or readily determined. Because of a self-protecting dose rate level of radiation (dose rate greater than 100 ren-x/h at I m in air), it is important to know the dose rate of spent fuel assemblies at all time. Estimates of the photon dose rate for spent MTR, TRIGA and DIDO-type fuel assemblies are given in this report

  1. Feasibility study of radiophotoluminescent glass rod dosimeter postal dose intercomparison for high energy photon beam

    International Nuclear Information System (INIS)

    Rah, Jeong-Eun; Kim, Siyong; Cheong, Kwang-Ho; Lee, Jeong-Woo; Chung, Jin-Beom; Shin, Dong-Oh; Suh, Tae-Suk

    2009-01-01

    A radiophotoluminescent glass rod dosimeter (GRD) system has recently become commercially available. In this study we evaluated whether the GRD would be suitable for external dosimetric audit program in radiotherapy. For this purpose, we introduced a methodology of the absorbed dose determination with the GRD by establishing calibration coefficient and various correction factors (non-linearity dose response, fading, energy dependence and angular dependence). A feasibility test of the GRD postal dose intercomparison was also performed for eight high photon beams by considering four radiotherapy centers in Korea. In the accuracy evaluation of the GRD dosimetry established in this study, we obtained within 1.5% agreements with the ionization chamber dosimetry for the 60 Co beam. It was also observed that, in the feasibility study, all the relative deviations were smaller than 3%. Based on these results, we believe that the new GRD system has considerable potential to be used for a postal dose audit program

  2. A study of dose equivalent for the nurses in Hirosaki University Hospital

    International Nuclear Information System (INIS)

    Kon, Masanori; Abe, Katsumi

    2001-01-01

    The annual relationships in 1997-1999 between exposure dose of those nurses engaging in full-time radiological works and the number of patients subjected to radiological examinations were investigated in authors' hospital. The annual number of those patients was rather constant. Exposure was measured by film-, carrot- and ring-badges. Eight to nine nurses engaged in radiological examinations like CT, fluoroscopy, urinary tract fluoroscopy and angiography, and other 8 nurses, therapy with sealed 137 Cs and 192 Ir and unsealed 131 I as well. No significant changes in exposure dose were observed in the former group of nurses and in the latter. The dose decreased annually to the level of the former due to skill advancement. (K.H.)

  3. An Analytical Model of Leakage Neutron Equivalent Dose for Passively-Scattered Proton Radiotherapy and Validation with Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Schneider, Christopher; Newhauser, Wayne, E-mail: newhauser@lsu.edu [Department of Physics and Astronomy, Louisiana State University and Agricultural and Mechanical College, 202 Nicholson Hall, Baton Rouge, LA 70803 (United States); Mary Bird Perkins Cancer Center, 4950 Essen Lane, Baton Rouge, LA 70809 (United States); Farah, Jad [Institut de Radioprotection et de Sûreté Nucléaire, Service de Dosimétrie Externe, BP-17, 92262 Fontenay-aux-Roses (France)

    2015-05-18

    Exposure to stray neutrons increases the risk of second cancer development after proton therapy. Previously reported analytical models of this exposure were difficult to configure and had not been investigated below 100 MeV proton energy. The purposes of this study were to test an analytical model of neutron equivalent dose per therapeutic absorbed dose (H/D) at 75 MeV and to improve the model by reducing the number of configuration parameters and making it continuous in proton energy from 100 to 250 MeV. To develop the analytical model, we used previously published H/D values in water from Monte Carlo simulations of a general-purpose beamline for proton energies from 100 to 250 MeV. We also configured and tested the model on in-air neutron equivalent doses measured for a 75 MeV ocular beamline. Predicted H/D values from the analytical model and Monte Carlo agreed well from 100 to 250 MeV (10% average difference). Predicted H/D values from the analytical model also agreed well with measurements at 75 MeV (15% average difference). The results indicate that analytical models can give fast, reliable calculations of neutron exposure after proton therapy. This ability is absent in treatment planning systems but vital to second cancer risk estimation.

  4. An Analytical Model of Leakage Neutron Equivalent Dose for Passively-Scattered Proton Radiotherapy and Validation with Measurements

    International Nuclear Information System (INIS)

    Schneider, Christopher; Newhauser, Wayne; Farah, Jad

    2015-01-01

    Exposure to stray neutrons increases the risk of second cancer development after proton therapy. Previously reported analytical models of this exposure were difficult to configure and had not been investigated below 100 MeV proton energy. The purposes of this study were to test an analytical model of neutron equivalent dose per therapeutic absorbed dose (H/D) at 75 MeV and to improve the model by reducing the number of configuration parameters and making it continuous in proton energy from 100 to 250 MeV. To develop the analytical model, we used previously published H/D values in water from Monte Carlo simulations of a general-purpose beamline for proton energies from 100 to 250 MeV. We also configured and tested the model on in-air neutron equivalent doses measured for a 75 MeV ocular beamline. Predicted H/D values from the analytical model and Monte Carlo agreed well from 100 to 250 MeV (10% average difference). Predicted H/D values from the analytical model also agreed well with measurements at 75 MeV (15% average difference). The results indicate that analytical models can give fast, reliable calculations of neutron exposure after proton therapy. This ability is absent in treatment planning systems but vital to second cancer risk estimation

  5. An analytical model of leakage neutron equivalent dose for passively-scattered proton radiotherapy and validation with measurements.

    Science.gov (United States)

    Schneider, Christopher; Newhauser, Wayne; Farah, Jad

    2015-05-18

    Exposure to stray neutrons increases the risk of second cancer development after proton therapy. Previously reported analytical models of this exposure were difficult to configure and had not been investigated below 100 MeV proton energy. The purposes of this study were to test an analytical model of neutron equivalent dose per therapeutic absorbed dose  at 75 MeV and to improve the model by reducing the number of configuration parameters and making it continuous in proton energy from 100 to 250 MeV. To develop the analytical model, we used previously published H/D values in water from Monte Carlo simulations of a general-purpose beamline for proton energies from 100 to 250 MeV. We also configured and tested the model on in-air neutron equivalent doses measured for a 75 MeV ocular beamline. Predicted H/D values from the analytical model and Monte Carlo agreed well from 100 to 250 MeV (10% average difference). Predicted H/D values from the analytical model also agreed well with measurements at 75 MeV (15% average difference). The results indicate that analytical models can give fast, reliable calculations of neutron exposure after proton therapy. This ability is absent in treatment planning systems but vital to second cancer risk estimation.

  6. The new remcounter LB6411: Measurement of neutron ambient dose equivalent H*(10) according to ICRP60 with high sensitivity

    International Nuclear Information System (INIS)

    Klett, A.; Burgkhardt, B.

    1996-01-01

    Since the International Commission on Radiological Protection has issued in publication ICRP60 new recommendations on radiation protection quantities, in neutron monitoring there is now increasing Interest in commercially available instruments optimized and calibrated for the measurement of ambient dose equivalent H*(10). Therefore within a joint cooperation between the Research Center Karlsruhe and EG ampersand G Berthold the neutron-dose-rate meter LB6411 was newly developed. The detector system with integrated electronics has a 3 He proportional counter tube centered in a moderating sphere. The response between thermal energies and 20 MeV was optimized with the help of extensive MCNP Monte-Carlo calculations. The instrument has extremely high sensitivity of approximately 3 counts per nSv and can be used both as a portable or as a stationary neutron monitor. Fluence responses and angular dependencies had been measured in monoenergetic neutron beams provided by the Physikalisch-Technische Bundesanstalt (PTB) in Braunschweig, Germany. The ambient dose equivalent response of the LB6411 is reported over the whole energy range

  7. Thermal neutron equivalent doses assessment around KFUPM neutron source storage area using NTDs

    Energy Technology Data Exchange (ETDEWEB)

    Abu-Jarad, F.; Fazal-ur-Rehman; Al-Haddad, M.N.; Al-Jarrallah, M.I.; Nassar, R

    2002-07-01

    Area passive neutron dosemeters based on nuclear track detectors (NTDs) have been used for 13 days to assess accumulated low doses of thermal neutrons around neutron source storage area of the King Fahd University of Petroleum and Minerals (KFUPM). Moreover, the aim of this study is to check the effectiveness of shielding of the storage area. NTDs were mounted with the boron converter on their surface as one compressed unit. The converter is a lithium tetraborate (Li{sub 2}B{sub 4}O{sub 7}) layer for thermal neutron detection via {sup 10}B(N,{alpha}){sup 7}Li and {sup 6}Li(n,{alpha}){sup 3}H nuclear reactions. The area passive dosemeters were installed on 26 different locations around the source storage area and adjacent rooms. The calibration factor for NTD-based area passive neutron dosemeters was found to be 8.3 alpha tracks.cm{sup -2}.{mu}Sv{sup -1} using active snoopy neutron dosemeters in the KFUPM neutron irradiation facility. The results show the variation of accumulated dose with locations around the storage area. The range of dose rates varied from as low as 40 nSv.h{sup -1} up to 11 {mu}Sv.h{sup -1}. The study indicates that the area passive neutron dosemeter was able to detect accumulated doses as low as 40 nSv.h{sup -1}, which could not be detected with the available active neutron dosemeters. The results of the study also indicate that an additional shielding is required to bring the dose rates down to background level. The present investigation suggests extending this study to find the contribution of doses from fast neutrons around the neutron source storage area using NTDs through proton recoil. The significance of this passive technique is that it is highly sensitive and does not require any electronics or power supplies, as is the case in active systems. (author)

  8. The risk equivalent of an exposure to-, versus a dose of radiation

    International Nuclear Information System (INIS)

    Bond, V.P.

    1986-01-01

    The long-term potential carcinogenic effects of low-level exposure (LLE) are addressed. The principal point discussed is linear, no-threshold dose-response curve. That the linear no-threshold, or proportional relationship is widely used is seen in the way in which the values for cancer risk coefficients are expressed - in terms of new cases, per million persons exposed, per year, per unit exposure or dose. This implies that the underlying relationship is proportional, i.e., ''linear, without threshold''. 12 refs., 9 figs., 1 tab

  9. Photon dose estimation from ultraintense laser–solid interactions and shielding calculation with Monte Carlo simulation

    International Nuclear Information System (INIS)

    Yang, Bo; Qiu, Rui; Li, JunLi; Lu, Wei; Wu, Zhen; Li, Chunyan

    2017-01-01

    When a strong laser beam irradiates a solid target, a hot plasma is produced and high-energy electrons are usually generated (the so-called “hot electrons”). These energetic electrons subsequently generate hard X-rays in the solid target through the Bremsstrahlung process. To date, only limited studies have been conducted on this laser-induced radiological protection issue. In this study, extensive literature reviews on the physics and properties of hot electrons have been conducted. On the basis of these information, the photon dose generated by the interaction between hot electrons and a solid target was simulated with the Monte Carlo code FLUKA. With some reasonable assumptions, the calculated dose can be regarded as the upper boundary of the experimental results over the laser intensity ranging from 10 19 to 10 21 W/cm 2 . Furthermore, an equation to estimate the photon dose generated from ultraintense laser–solid interactions based on the normalized laser intensity is derived. The shielding effects of common materials including concrete and lead were also studied for the laser-driven X-ray source. The dose transmission curves and tenth-value layers (TVLs) in concrete and lead were calculated through Monte Carlo simulations. These results could be used to perform a preliminary and fast radiation safety assessment for the X-rays generated from ultraintense laser–solid interactions. - Highlights: • The laser–driven X-ray ionizing radiation source was analyzed in this study. • An equation to estimate the photon dose based on the laser intensity is given. • The shielding effects of concrete and lead were studied for this new X-ray source. • The aim of this study is to analyze and mitigate the laser–driven X-ray hazard.

  10. Quantitative radiation dose-response relationships for normal tissues in man - I. Gustatory tissues response during photon and neutron radiotherapy

    International Nuclear Information System (INIS)

    Mossman, K.L.

    1982-01-01

    Quantitative radiation dose-response curves for normal gustatory tissue in man were studied. Taste function, expressed as taste loss, was evaluated in 84 patients who were given either photon or neutron radiotherapy for tumors in the head and neck region. Patients were treated to average tumor doses of 6600 cGy (photon) or 2200 cGy intervals for photon patients and 320-cGy intervals for neutron patients during radiotherapy. The dose-response curves for photons and neutrons were analyzed by fitting a four-parameter logistic equation to the data. Photon and neutron curves differed principally in their relative position along the dose axis. Comparison of the dose-response curves were made by determination of RBE. At 320 cGy, the lowest neutron dose at which taste measurements were made, RBE = 5.7. If this RBE is correct, then the therapeutic gain factor may be equal to or less than 1, indicating no biological advantage in using neutrons over photons for this normal tissue. These studies suggest measurements of taste function and evaluation of dose-response relationships may also be useful in quantitatively evaluating the efficacy of chemical modifiers of radiation response such as hypoxic cell radiosensitizers and radioprotectors

  11. Radiation brain dose to vascular surgeons during fluoroscopically guided interventions is not effectively reduced by wearing lead equivalent surgical caps.

    Science.gov (United States)

    Kirkwood, Melissa L; Arbique, Gary M; Guild, Jeffrey B; Zeng, Katie; Xi, Yin; Rectenwald, John; Anderson, Jon A; Timaran, Carlos

    2018-03-12

    Radiation to the interventionalist's brain during fluoroscopically guided interventions (FGIs) may increase the incidence of cerebral neoplasms. Lead equivalent surgical caps claim to reduce radiation brain doses by 50% to 95%. We sought to determine the efficacy of the RADPAD (Worldwide Innovations & Technologies, Lenexa, Kan) No Brainer surgical cap (0.06 mm lead equivalent at 90 kVp) in reducing radiation dose to the surgeon's and trainee's head during FGIs and to a phantom to determine relative brain dose reductions. Optically stimulated, luminescent nanoDot detectors (Landauer, Glenwood, Ill) inside and outside of the cap at the left temporal position were used to measure cap attenuation during FGIs. To check relative brain doses, nanoDot detectors were placed in 15 positions within an anthropomorphic head phantom (ATOM model 701; CIRS, Norfolk, Va). The phantom was positioned to represent a primary operator performing femoral access. Fluorography was performed on a plastic scatter phantom at 80 kVp for an exposure of 5 Gy reference air kerma with or without the hat. For each brain location, the percentage dose reduction with the hat was calculated. Means and standard errors were calculated using a pooled linear mixed model with repeated measurements. Anatomically similar locations were combined into five groups: upper brain, upper skull, midbrain, eyes, and left temporal position. This was a prospective, single-center study that included 29 endovascular aortic aneurysm procedures. The average procedure reference air kerma was 2.6 Gy. The hat attenuation at the temporal position for the attending physician and fellow was 60% ± 20% and 33% ± 36%, respectively. The equivalent phantom measurements demonstrated an attenuation of 71% ± 2.0% (P < .0001). In the interior phantom locations, attenuation was statistically significant for the skull (6% ± 1.4%) and upper brain (7.2% ± 1.0%; P < .0001) but not for the middle brain (1.4% ± 1.0%; P = .15

  12. Using two detectors concurrently to monitor ambient dose equivalent rates in vehicle surveys of radiocesium contaminated land.

    Science.gov (United States)

    Takeishi, Minoru; Shibamichi, Masaru; Malins, Alex; Kurikami, Hiroshi; Murakami, Mitsuhiro; Saegusa, Jun; Yoneya, Masayuki

    2017-10-01

    In response to the accident at Tokyo Electric Power Company's Fukushima Dai-ichi Nuclear Power Plant (FDNPP), vehicle-borne monitoring was used to map radiation levels for radiological protection of the public. By convention measurements from vehicle-borne surveys are converted to the ambient dose equivalent rate at 1 m height in the absence of the vehicle. This allows for comparison with results from other types of survey, including surveys with hand-held or airborne instruments. To improve the accuracy of the converted results from vehicle-borne surveys, we investigated combining measurements from two detectors mounted on the vehicle at different heights above the ground. A dual-detector setup was added to a JAEA monitoring car and compared against hand-held survey meter measurements in Fukushima Prefecture. The results obtained by combining measurements from two detectors were within ±20% of the hand-held reference measurements. The mean absolute percentage deviation from the reference measurements was 7.2%. The combined results from the two detectors were more accurate than those from either the roof-mounted detector, or the detector inside the vehicle, taken alone. One issue with vehicle-borne surveys is that ambient dose equivalent rates above roads are not necessarily representative of adjacent areas. This is because radiocesium is often deficient on asphalt surfaces, as it is easily scrubbed off by rain, wind and vehicle tires. To tackle this issue, we investigated mounting heights for vehicle-borne detectors using Monte Carlo gamma-ray simulations. When radiocesium is deficient on a road compared to the adjacent land, mounting detectors high on vehicles yields results closer to the values adjacent to the road. The ratio of ambient dose equivalent rates reported by detectors mounted at different heights in a dual-detector setup indicates whether radiocesium is deficient on the road compared to the adjacent land. Copyright © 2017 Elsevier Ltd. All rights

  13. Biologically-equivalent dose and long-term survival time in radiation treatments

    International Nuclear Information System (INIS)

    Zaider, Marco; Hanin, Leonid

    2007-01-01

    Within the linear-quadratic model the biologically-effective dose (BED)-taken to represent treatments with an equal tumor control probability (TCP)-is commonly (and plausibly) calculated according to BED(D) = -log[S(D)]/α. We ask whether in the presence of cellular proliferation this claim is justified and examine, as a related question, the extent to which BED approximates an isoeffective dose (IED) defined, more sensibly, in terms of an equal long-term survival probability, rather than TCP. We derive, under the assumption that cellular birth and death rates are time homogeneous, exact equations for the isoeffective dose, IED. As well, we give a rigorous definition of effective long-term survival time, T eff . By using several sets of radiobiological parameters, we illustrate potential differences between BED and IED on the one hand and, on the other, between T eff calculated as suggested here or by an earlier recipe. In summary: (a) the equations currently in use for calculating the effective treatment time may underestimate the isoeffective dose and should be avoided. The same is the case for the tumor control probability (TCP), only more so; (b) for permanent implants BED may be a poor substitute for IED; (c) for a fractionated treatment schedule, interpreting the observed probability of cure in terms of a TCP formalism that refers to the end of the treatment (rather than T eff ) may result in a miscalculation (underestimation) of the initial number of clonogens

  14. Effect of dosimeter type for commissioning small photon beams on calculated dose distribution in stereotactic radiosurgery

    Energy Technology Data Exchange (ETDEWEB)

    García-Garduño, O. A., E-mail: oagarciag@innn.edu.mx, E-mail: amanda.garcia.g@gmail.com [Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Mexico City 14269, México and Centro de Investigación en Ciencia Aplicada y Tecnología Avanzada, Unidad Legaria, Instituto Politécnico Nacional, Legaria 694, México City 11500, México (Mexico); Rodríguez-Ponce, M. [Departamento de Biofísica, Instituto Nacional de Cancerología, Mexico City 14080, México (Mexico); Gamboa-deBuen, I. [Instituto de Ciencias Nucleares, Universidad Nacional Autónoma de México, Ciudad Universitaria, Mexico City 04510 (Mexico); Rodríguez-Villafuerte, M. [Instituto de Física, Universidad Nacional Autónoma de México, Ciudad Universitaria, Mexico City 04510 (Mexico); Galván de la Cruz, O. O. [Laboratorio de Física Médica, Instituto Nacional de Neurología y Neurocirugía, Mexico City 14269, México (Mexico); and others

    2014-09-15

    Purpose: To assess the impact of the detector used to commission small photon beams on the calculated dose distribution in stereotactic radiosurgery (SRS). Methods: In this study, six types of detectors were used to characterize small photon beams: three diodes [a silicon stereotactic field diode SFD, a silicon diode SRS, and a silicon diode E], an ionization chamber CC01, and two types of radiochromic film models EBT and EBT2. These detectors were used to characterize circular collimated beams that were generated by a Novalis linear accelerator. This study was conducted in two parts. First, the following dosimetric data, which are of particular interest in SRS, were compared for the different detectors: the total scatter factor (TSF), the tissue phantom ratios (TPRs), and the off-axis ratios (OARs). Second, the commissioned data sets were incorporated into the treatment planning system (TPS) to compare the calculated dose distributions and the dose volume histograms (DVHs) that were obtained using the different detectors. Results: The TSFs data measured by all of the detectors were in good agreement with each other within the respective statistical uncertainties: two exceptions, where the data were systematically below those obtained for the other detectors, were the CC01 results for all of the circular collimators and the EBT2 film results for circular collimators with diameters below 10.0 mm. The OAR results obtained for all of the detectors were in excellent agreement for all of the circular collimators. This observation was supported by the gamma-index test. The largest difference in the TPR data was found for the 4.0 mm circular collimator, followed by the 10.0 and 20.0 mm circular collimators. The results for the calculated dose distributions showed that all of the detectors passed the gamma-index test at 100% for the 3 mm/3% criteria. The aforementioned observation was true regardless of the size of the calculation grid for all of the circular collimators

  15. The absorbed dose in air of photons generated from secondary cosmic rays at sea level at Nagoya, Japan

    International Nuclear Information System (INIS)

    Akhmad, Y.R.

    1995-01-01

    Investigations have been carried out to determine the absorbed dose in air of photons generated from secondary cosmic radiation at sea level at Nagoya, Japan. To isolate the contribution from cosmic photons, the pulse-height distributions due to μ particles and electrons were eliminated from the observed pulse-height distribution of a measurement with a 3'' diam. spherical NaI(Tl) detector. The pulse height due to μ particles and electrons was inferred from the coincidence technique using two types of scintillation detectors with different sensitivities to photons. To obtain the photon fluence rate for further dose calculation, the pulse-height distribution of cosmic photons was unfolded by the iterative method. The mean and its standard deviation of the absorbed dose in air and fluence rate due to cosmic photons calculated from a one year observation are 2.86±0.05 nGy.h -1 and 0.1342±0.0015 photons.cm -2 .s -1 , respectively. The absorbed dose in air from cosmic photons was 0.5% lower during autumn to winter and 0.6% higher during spring to summer than the mean taken over the year. (author)

  16. The cyclophosphamide equivalent dose as an approach for quantifying alkylating agent exposure: a report from the Childhood Cancer Survivor Study.

    Science.gov (United States)

    Green, Daniel M; Nolan, Vikki G; Goodman, Pamela J; Whitton, John A; Srivastava, DeoKumar; Leisenring, Wendy M; Neglia, Joseph P; Sklar, Charles A; Kaste, Sue C; Hudson, Melissa M; Diller, Lisa R; Stovall, Marilyn; Donaldson, Sarah S; Robison, Leslie L

    2014-01-01

    Estimation of the risk of adverse long-term outcomes such as second malignant neoplasms and infertility often requires reproducible quantification of exposures. The method for quantification should be easily utilized and valid across different study populations. The widely used Alkylating Agent Dose (AAD) score is derived from the drug dose distribution of the study population and thus cannot be used for comparisons across populations as each will have a unique distribution of drug doses. We compared the performance of the Cyclophosphamide Equivalent Dose (CED), a unit for quantifying alkylating agent exposure independent of study population, to the AAD. Comparisons included associations from three Childhood Cancer Survivor Study (CCSS) outcome analyses, receiver operator characteristic (ROC) curves and goodness of fit based on the Akaike's Information Criterion (AIC). The CED and AAD performed essentially identically in analyses of risk for pregnancy among the partners of male CCSS participants, risk for adverse dental outcomes among all CCSS participants and risk for premature menopause among female CCSS participants, based on similar associations, lack of statistically significant differences between the areas under the ROC curves and similar model fit values for the AIC between models including the two measures of exposure. The CED is easily calculated, facilitating its use for patient counseling. It is independent of the drug dose distribution of a particular patient population, a characteristic that will allow direct comparisons of outcomes among epidemiological cohorts. We recommend the use of the CED in future research assessing cumulative alkylating agent exposure. © 2013 Wiley Periodicals, Inc.

  17. Degradation of proton depth dose distributions attributable to microstructures in lung-equivalent material

    Energy Technology Data Exchange (ETDEWEB)

    Titt, Uwe, E-mail: utitt@mdanderson.org; Mirkovic, Dragan; Mohan, Radhe [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 (United States); Sell, Martin [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston, Texas 77030 and Department of Medical Physics, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120 (Germany); Unkelbach, Jan [Department of Radiation Oncology, Massachusetts General Hospital, 55 Fruit Street, Boston, Massachusetts 02114 (United States); Bangert, Mark [Department of Medical Physics, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120 (Germany); Oelfke, Uwe [Department of Medical Physics, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120, Germany and Department of Physics, The Institute of Cancer Research, 123 Old Brompton Road, London SW7 3RP (United Kingdom)

    2015-11-15

    Purpose: The purpose of the work reported here was to investigate the influence of sub-millimeter size heterogeneities on the degradation of the distal edges of proton beams and to validate Monte Carlo (MC) methods’ ability to correctly predict such degradation. Methods: A custom-designed high-resolution plastic phantom approximating highly heterogeneous, lung-like structures was employed in measurements and in Monte Carlo simulations to evaluate the degradation of proton Bragg curves penetrating heterogeneous media. Results: Significant differences in distal falloff widths and in peak dose values were observed in the measured and the Monte Carlo simulated curves compared to pristine proton Bragg curves. Furthermore, differences between simulations of beams penetrating CT images of the phantom did not agree well with the corresponding experimental differences. The distal falloff widths in CT image-based geometries were underestimated by up to 0.2 cm in water (corresponding to 0.8–1.4 cm in lung tissue), and the peak dose values of pristine proton beams were overestimated by as much as ~35% compared to measured curves or depth-dose curves simulated on the basis of true geometry. The authors demonstrate that these discrepancies were caused by the limited spatial resolution of CT images that served as a basis for dose calculations and lead to underestimation of the impact of the fine structure of tissue heterogeneities. A convolution model was successfully applied to mitigate the underestimation. Conclusions: The results of this study justify further development of models to better represent heterogeneity effects in soft-tissue geometries, such as lung, and to correct systematic underestimation of the degradation of the distal edge of proton doses.

  18. Degradation of proton depth dose distributions attributable to microstructures in lung-equivalent material

    International Nuclear Information System (INIS)

    Titt, Uwe; Mirkovic, Dragan; Mohan, Radhe; Sell, Martin; Unkelbach, Jan; Bangert, Mark; Oelfke, Uwe

    2015-01-01

    Purpose: The purpose of the work reported here was to investigate the influence of sub-millimeter size heterogeneities on the degradation of the distal edges of proton beams and to validate Monte Carlo (MC) methods’ ability to correctly predict such degradation. Methods: A custom-designed high-resolution plastic phantom approximating highly heterogeneous, lung-like structures was employed in measurements and in Monte Carlo simulations to evaluate the degradation of proton Bragg curves penetrating heterogeneous media. Results: Significant differences in distal falloff widths and in peak dose values were observed in the measured and the Monte Carlo simulated curves compared to pristine proton Bragg curves. Furthermore, differences between simulations of beams penetrating CT images of the phantom did not agree well with the corresponding experimental differences. The distal falloff widths in CT image-based geometries were underestimated by up to 0.2 cm in water (corresponding to 0.8–1.4 cm in lung tissue), and the peak dose values of pristine proton beams were overestimated by as much as ~35% compared to measured curves or depth-dose curves simulated on the basis of true geometry. The authors demonstrate that these discrepancies were caused by the limited spatial resolution of CT images that served as a basis for dose calculations and lead to underestimation of the impact of the fine structure of tissue heterogeneities. A convolution model was successfully applied to mitigate the underestimation. Conclusions: The results of this study justify further development of models to better represent heterogeneity effects in soft-tissue geometries, such as lung, and to correct systematic underestimation of the degradation of the distal edge of proton doses

  19. ORION: a computer code for evaluating environmental concentrations and dose equivalent to human organs or tissue from airborne radionuclides

    International Nuclear Information System (INIS)

    Shinohara, K.; Nomura, T.; Iwai, M.

    1983-05-01

    The computer code ORION has been developed to evaluate the environmental concentrations and the dose equivalent to human organs or tissue from air-borne radionuclides released from multiple nuclear installations. The modified Gaussian plume model is applied to calculate the dispersion of the radionuclide. Gravitational settling, dry deposition, precipitation scavenging and radioactive decay are considered to be the causes of depletion and deposition on the ground or on vegetation. ORION is written in the FORTRAN IV language and can be run on IBM 360, 370, 303X, 43XX and FACOM M-series computers. 8 references, 6 tables

  20. Estimate of the Effective Dose Equivalent to the Cypriot Population due to Diagnostic Nuclear Medicine Procedures in the Public Sector

    Energy Technology Data Exchange (ETDEWEB)

    Christofides, S [Medical Physics Department, Nicosia General Hospital (Cyprus)

    1994-12-31

    The Effective Dose Equivalent (EDE) to the Cypriot population due to Diagnostic Nuclear Medicine procedures has been estimated from data published by the Government of Cyprus, in its Health and Hospital Statistics Series for the years 1990, 1991, and 1992. The average EDE per patient was estimated to be 3,09, 3,75 and 4,01 microSievert for 1990, 1991 and 1992 respectively, while the per caput EDE was estimated to be 11,75, 15,16 and 17,09 microSieverts for 1990, 1991 and 1992 respectively, from the procedures in the public sector. (author). 11 refs, 4 tabs.

  1. Estimate of the Effective Dose Equivalent to the Cypriot Population due to Diagnostic Nuclear Medicine Procedures in the Public Sector

    International Nuclear Information System (INIS)

    Christofides, S.

    1994-01-01

    The Effective Dose Equivalent (EDE) to the Cypriot population due to Diagnostic Nuclear Medicine procedures has been estimated from data published by the Government of Cyprus, in its Health and Hospital Statistics Series for the years 1990, 1991, and 1992. The average EDE per patient was estimated to be 3,09, 3,75 and 4,01 microSievert for 1990, 1991 and 1992 respectively, while the per caput EDE was estimated to be 11,75, 15,16 and 17,09 microSieverts for 1990, 1991 and 1992 respectively, from the procedures in the public sector. (author)

  2. Evaluation of equivalent and effective dose by KAP for patient and orthopedic surgeon in vertebral compression fracture surgery

    International Nuclear Information System (INIS)

    Santos, Felipe A.; Galeano, Diego C.; Santos, William S.; Silva, Ademir X.; Souza, Susana O.; Carvalho Júnior, Albérico B.

    2017-01-01

    Clinical scenarios were virtually modeled to estimate both the equivalent and effective doses normalized by KAP (Kerma Area Product) to vertebra compression fracture surgery in patient and surgeon. This surgery is known as kyphoplasty and involves the use of X-ray equipment, the C-arm, which provides real-time images to assist the surgeon in conducting instruments inserted into the patient and in the delivery of surgical cement into the fractured vertebra. The radiation transport code used was MCNPX (Monte Carlo N-Particle eXtended) and a pair of UFHADM (University of Florida Hybrid ADult Male) virtual phantoms. The developed scenarios allowed us to calculate a set of equivalent dose (H T ) and effective dose (E) for patients and surgeons. In additional, the same scenario was calculated KAP in the tube output and was used for calculating conversion coefficients (E/KAP and H T /KAP). From the knowledge of the experimental values of KAP and the results presented in this study, it is possible to estimate absolute values of effective doses for different exposure conditions. In this work, we developed scenarios with and without the surgical table with the purpose of comparison with the existing data in the literature. The absence of the bed in the scenario promoted a percentage absolute difference of 56% in the patient effective doses in relation to scenarios calculated with a bed. Regarding the surgeon, the use of the personal protective equipment (PPE) reduces between 75% and 79% the effective dose and the use of the under table shield (UTS) reduces the effective dose of between 3% and 7%. All these variations emphasize the importance of the elaboration of virtual scenarios that approach the actual clinical conditions generating E/KAP and H T /KAP closer to the actual values. - Highlights: • Virtual scenarios of vertebra compression fracture surgery. • MC simulations using virtual anthropomorphic phantoms and surgical setups. • Estimation of E/KAP and H T /KAP

  3. Review of personal monitoring techniques for the measurement of absorbed dose from external beta and low energy photon radiation

    DEFF Research Database (Denmark)

    Christensen, Poul

    1986-01-01

    The techniques available at present for personal monitoring of doses from external beta and low energy photon radiation are reviewed. The performance of currently used dosimetry systems is compared with that recommended internationally, and developments for improving the actual performance...

  4. Radiation dose response of N channel MOSFET submitted to filtered X-ray photon beam

    Science.gov (United States)

    Gonçalves Filho, Luiz C.; Monte, David S.; Barros, Fabio R.; Santos, Luiz A. P.

    2018-01-01

    MOSFET can operate as a radiation detector mainly in high-energy photon beams, which are normally used in cancer treatments. In general, such an electronic device can work as a dosimeter from threshold voltage shift measurements. The purpose of this article is to show a new way for measuring the dose-response of MOSFETs when they are under X-ray beams generated from 100kV potential range, which is normally used in diagnostic radiology. Basically, the method consists of measuring the MOSFET drain current as a function of the radiation dose. For this the type of device, it has to be biased with a high value resistor aiming to see a substantial change in the drain current after it has been irradiated with an amount of radiation dose. Two types of N channel device were used in the experiment: a signal transistor and a power transistor. The delivered dose to the device was varied and the electrical curves were plotted. Also, a sensitivity analysis of the power MOSFET response was made, by varying the tube potential of about 20%. The results show that both types of devices have responses very similar, the shift in the electrical curve is proportional to the radiation dose. Unlike the power MOSFET, the signal transistor does not provide a linear function between the dose rate and its drain current. We also have observed that the variation in the tube potential of the X-ray equipment produces a very similar dose-response.

  5. Muscle and plastic equivalent glass dosimeter for high-dose dosimetry

    International Nuclear Information System (INIS)

    Abdel-Rehim, F.; Maged, A.F.; Morsy, M.A.; Hashad, A.M.

    1990-01-01

    The alkali-silicate glass dosimeter is made up of 66.8% SiO 2 , 31.2% Li 2 O, 2% K 2 O. It is nearly colourless before irradiation and then takes on an amber colour with increasing doses of gamma radiation. This colouration is represented by the appearance of broad absorption bands at 405 nm and 600 nm wavelengths. The change in absorbance is linear with the absorbed dose in the range of 0.1-4.5 kGy, when measured at its 405 nm absorption band maximum. This glass dosimeter simulates low-z plastics and muscle tissue in terms of gamma-ray absorption properties over broad radiation spectra (0.1 MeV to 10 MeV). (author) 22 refs.; 4 figs.; 2 tabs

  6. SU-F-I-05: Dose Symmetry for CTDI Equivalent Measurements with Limited Angle CBCT

    Energy Technology Data Exchange (ETDEWEB)

    Singh, V [Henry Ford Hospital, Detroit, MI (United States); McKenney, S [Children’s National Medical Center, Washington, DC (United States); Sunde, P [Radcal, Inc, Monrovia, CA (United States); Feng, W [New York Presbyterian Hospital, Tenafly, NJ (United States); Bakalyar, D [Henry Ford Health System, Detroit, MI (United States)

    2016-06-15

    Purpose: CTDI measurements, useful for characterizing the x-ray output for multi-detector CT (MDCT), require a 360° rotation of the gantry; this presents a problem for cone beam CT (CBCT) due to its limited angular rotation. The purpose of this work is to demonstrate a methodology for overcoming this limited angular rotation so that CTDI measurements can also be made on CBCT systems making it possible to compare the radiation output from both types of system with a common metric. Methods: The symmetry of the CTDI phantom allows a 360° CTDI measurement to be replaced with two 180° measurements. A pencil chamber with a real-time digitizer was placed at the center of the head phantom (16 cm, PMMA) and the resulting exposure measurement from a 180° acquisition was doubled. A pair of edge measurements, each obtained with the gantry passing through the same 180 arc, was obtained with the pencil chamber at opposite edges of the diameter of the phantom and then summed. The method was demonstrated on a clinical CT scanner (Philips, Brilliance6) and then implemented on an interventional system (Siemens, Axiom Artis). Results: The equivalent CTDI measurement agreed with the conventional CTDI measurement within 8%. The discrepancy in the two measurements is largely attributed to uncertainties in cropping the waveform to a 180°acquisition. (Note: Because of the reduced fan angle in the CBCT, CTDI is not directly comparable to MDCT values when a 32 cm phantom is used.) Conclusion: The symmetry-based CTDI measurement is an equivalent measurement to the conventional CTDI measurement when the fan angle is large enough to encompass the phantom diameter. This allows a familiar metric of radiation output to be employed on systems with a limited angular rotation.

  7. Workplace characterization using spectrometry for reliable neutron dose assessment in mixed neutron photon fields

    International Nuclear Information System (INIS)

    Bolognese-Milsztajn, T.; Asselineau, B.; Itie, C.; Lahaye, T.; Muller, H.

    2002-01-01

    Radiation protection for practices should be based on justification of exposure, optimization of protection and dose limitations. The 96/29 European Union directive clearly requires that operational protection of exposed workers shall be based on prior evaluation to identify the nature and magnitude of the radiological risk to exposed workers and implementation of optimization of radiation protection in all working conditions. Operational protection of exposed workers requires the implementation of measures at workplaces. The ICRP60 publication (1991) defines and recommends the use of the effective dose E for radiological protections purposes. This quantity can not be directly measured; the operational quantities ambient dose equivalent: H and personal dose equivalent: Hp as defined in the ICRU 51 report (1993) are recommended to estimate the effective dose E. These operational quantities are related, through calculated conversion coefficients h Φ , to the measurable function fluence Φ by: H = ∫ h Φ (E) Φ E dE (1a); Hp = ∫ hp Φ (E, Ω) Φ E , Ω dE dΩ (1b). Φ E (Φ E ,Ω) is the distribution of the fluence with respect to energy (energy E and direction Ω); h Φ or hp Φ are the fluence to operational quantity conversion coefficients as defined by the ICRU 57 report (1998)

  8. A new Monte Carlo program for calculations of dose distributions within tissue equivalent phantoms irradiated from π--meson beams

    International Nuclear Information System (INIS)

    Przybilla, G.

    1980-11-01

    The present paper reports on the structure and first results from a new Monte Carlo programme for calculations of energy distributions within tissue equivalent phantoms irradiated from π - -beams. Each pion or generated secondary particle is transported until to the complete loss of its kinetic energy taking into account pion processes like multiple Coulomb scattering, pion reactions in flight and absorption of stopped pions. The code uses mainly data from experiments, and physical models have been added only in cases of lacking data. Depth dose curves for a pensil beam of 170 MeV/c within a water phantom are discussed as a function of various parameters. Isodose contours are plotted resulting from a convolution of an extended beam profile and the dose distribution of a pencil beams. (orig.) [de

  9. Didactic revision of the operative magnitudes system ICRU for the evaluation of the equivalent dose in radiation external fields

    International Nuclear Information System (INIS)

    Alvarez R, J. T.

    2014-10-01

    In this work is presented in a didactic way the operative magnitudes system ICRU, showing as these magnitudes carry out an appropriate estimate of the effective equivalent doses H E and the effective dose. The objective is to present the basic concepts of the dosimetry for radiation external fields with purposes of radiological protection, because the assimilation lack and technological development of this dosimetric magnitudes system has persisted for near 50 years, in terms of practice of the radiological protection in Mexico. Also, this system is an essential part of safety basic standards of the IAEA and ICRP recommendations 26, 60, 74 and 103, as well as of the ICRU 25, 39, 43, 51 and 57. (Author)

  10. Determination of attenuation factors for mortar of barite in terms of environmental dose equivalent and effective dose

    International Nuclear Information System (INIS)

    Almeida Junior, Airton T.; Campos, L.L.R.; Araujo, F.G.S.; Santos, M.A.P.; Nogueira, M.S.

    2014-01-01

    This work addresses the characterization of barite mortars used as Xray shielding materials through the following quantities: mass attenuation coefficient, air kerma, effective dose and ambient dose - H⁎(10). The experiment was carried out with the use of the following reference qualities: RQR4, RQR6, RQR9 e RQR10, specified in accordance with norm IEC 61267: Medical diagnostic Xray equipment - radiation conditions for use in the determination of characteristics. In this study values was determined experimentally for the attenuation of the Cream barite (density 2.99g/cm 3 , collected in the state of Sao Paulo), Purple barite (density 2.95g/cm 3 , collected in the state of Bahia) and White barite (density 3.10g/cm 3 , collected in the state of Paraiba). These materials, in the form of mortar, were disposed in the form of squares namely poof bodies, whose dimensions were 10 x 10 cm and thickness ranging from 3 to 15 mm approximately. In the experimental procedure, these proof bodies were irradiated with a Pantak, model HF320 industrial X-ray apparatus. The potentials applied to the respective X-ray tube were: 60kV, 80kV, 120kV and 150kV at a constant current of 1mA. The attenuation responses in function of thickness, for each of the materials analyzed, were used to draw the attenuation and transmission curves. The efficiency of the barite studied concerning the capacity to attenuate X-ray radiation for X-ray beams ranging from 60 to 150 kV indicated

  11. Development and use of a fifteen year-old equivalent mathematical phantom for internal dose calculations

    International Nuclear Information System (INIS)

    Jones, R.M.; Poston, J.W.; Hwang, J.L.; Jones, T.D.; Warner, G.G.

    1976-06-01

    The existence of a phantom based on anatomical data for the average fifteen-year-old provides for a proficient means of obtaining estimates of absorbed dose for children of that age. Dimensions representative of an average fifteen-year-old human, obtained from various biological and medical research, were transformed into a mathematical construct of idealized shapes of the exterior, skeletal system, and internal organs of a human. The idealization for an average adult presently in use by the International Commission on Radiological Protection was used as a basis for design. The mathematical equations describing the phantom were developed to be readily adaptable to present-day methods of dose estimation. Typical exposure situations in nuclear medicine have previously been modeled for existing phantoms. With no further development of the exposure model necessary, adaptation to the fifteen-year-old phantom demonstrated the utility of the design. Estimates of absorbed dose were obtained for the administration of two radiopharmaceuticals, /sup 99m/Tc-sulfur colloid and /sup 99m/Tc-DMSA

  12. Measurement of conversion coefficients between air Kerma and personal dose equivalent and backscatter factors for diagnostic X-ray beams

    International Nuclear Information System (INIS)

    Rosado, Paulo Henrique Goncalves

    2008-01-01

    Two sets of quantities are import in radiological protection: the protection and operational quantities. Both sets can be related to basic physical quantities such as kerma through conversion coefficients. For diagnostic x-ray beams the conversion coefficients and backscatter factors have not been determined yet, those parameters are need for calibrating dosimeters that will be used to determine the personal dose equivalent or the entrance skin dose. Conversion coefficients between air kerma and personal dose equivalent and backscatter factors were experimentally determined for the diagnostic x-ray qualities RQR and RQA recommended by the International Electrotechnical Commission (IEC). The air kerma in the phantom and the mean energy of the spectrum were measured for such purpose. Harshaw LiF-100H thermoluminescent dosemeters (TLD) were used for measurements after being calibrated against an 180 cm 3 Radcal Corporation ionization chamber traceable to a reference laboratory. A 300 mm x 300 mm x 150 mm polymethylmethacrylate (PMMA) slab phantom was used for deep-dose measurements. Tl dosemeters were placed in the central axis of the x-ray beam at 5, 10, 15, 25 and 35 mm depth in the phantom upstream the beam direction Another required parameter for determining the conversion coefficients from was the mean energy of the x-ray spectrum. The spectroscopy of x-ray beams was done with a CdTe semiconductor detector that was calibrated with 133 Ba, 241 Am and 57 Co radiation sources. Measurements of the x-ray spectra were carried out for all RQR and RQA IEC qualities. Corrections due to the detector intrinsic efficiency, total energy absorption, escape fraction of the characteristic x-rays, Compton effect and attenuation in the detector were done aiming an the accurate determination of the mean energy. Measured x-ray spectra were corrected with the stripping method by using these response functions. The typical combined standard uncertainties of conversion coefficients and

  13. Volume dose ratios relevant for alanine dosimetry in small, 6 MV photon beams

    DEFF Research Database (Denmark)

    Cronholm, Rickard O.; Andersen, Claus Erik; Behrens, Claus F.

    2012-01-01

    therapy). To this end, we here present the results of a Monte Carlo simulation study with DOSRZnrc that investigated the influence of field and detector size for small 6 MV photon beams. The study focusses on doses averaged over the volume of the detector rather than point doses.The ratio of volume...... averaged doses to water (D¯W) and alanine (D¯det) was found to be approximately 1.025 for most situations studied, and a constant ratio is likely to be representative for many applications in radiation therapy. However, D¯W/D¯det was found to be as low as 0.9908 ± 0.0037 in situations where one might...... expect significant deviations from charged particle equilibrium (i.e. at shallow depths and when the field size was smaller than the range of the secondary electrons). These effects therefore need consideration when finite-size alanine dosimeters are used under such conditions....

  14. Effects of target size on the comparison of photon and charged particle dose distributions

    International Nuclear Information System (INIS)

    Phillips, M.H.; Frankel, K.A.; Tjoa, T.; Lyman, J.T.; Fabrikant, J.I.; Levy, R.P.

    1989-12-01

    The work presented here is part of an ongoing project to quantify and evaluate the differences in the use of different radiation types and irradiation geometries in radiosurgery. We are examining dose distributions for photons using the ''Gamma Knife'' and the linear accelerator arc methods, as well as different species of charged particles from protons to neon ions. A number of different factors need to be studied to accurately compare the different modalities such as target size, shape and location, the irradiation geometry, and biological response. This presentation focuses on target size, which has a large effect on the dose distributions in normal tissue surrounding the lesion. This work concentrates on dose distributions found in radiosurgery, as opposed to those usually found in radiotherapy. 5 refs., 2 figs

  15. Development of a Reference System for the determination of the personal dose equivalent and the constancy of X- Ray beams

    International Nuclear Information System (INIS)

    Vivolo, Vitor

    2006-01-01

    A reference system for the determination of the personal dose equivalent, Hp (10), and a quality control program of X-ray equipment used In radioprotection require the periodic verification of the X-ray beams constancy. In this work, two parallel-plate ionization chambers were developed with inner electrodes of different materials, and inserted into PMMA slab phantoms. One ionization chamber was developed with inner carbon electrodes and the other with inner aluminium electrodes. The two ionization chambers can be used as a Tandem system. The different energy response of the two ionization chambers allowed the development of the Tandem system that is very useful for the checking of the constancy of beam qualities. Standard intermediary energy X-ray beams (from 48 keV to 118 keV), radioprotection level, were established through the development of a dosimetric methodology and the analysis of their physical parameters. The ionization chambers were studied in relation to their operational characteristics, and they were calibrated in X-ray beams (radioprotection, diagnostic radiology, mammography and radiotherapy levels) in accordance to international recommendations. They presented good performance. The determination procedure of personal dose equivalent, Hp (10), was established. (author)

  16. A new formula for normal tissue complication probability (NTCP) as a function of equivalent uniform dose (EUD).

    Science.gov (United States)

    Luxton, Gary; Keall, Paul J; King, Christopher R

    2008-01-07

    To facilitate the use of biological outcome modeling for treatment planning, an exponential function is introduced as a simpler equivalent to the Lyman formula for calculating normal tissue complication probability (NTCP). The single parameter of the exponential function is chosen to reproduce the Lyman calculation to within approximately 0.3%, and thus enable easy conversion of data contained in empirical fits of Lyman parameters for organs at risk (OARs). Organ parameters for the new formula are given in terms of Lyman model m and TD(50), and conversely m and TD(50) are expressed in terms of the parameters of the new equation. The role of the Lyman volume-effect parameter n is unchanged from its role in the Lyman model. For a non-homogeneously irradiated OAR, an equation relates d(ref), n, v(eff) and the Niemierko equivalent uniform dose (EUD), where d(ref) and v(eff) are the reference dose and effective fractional volume of the Kutcher-Burman reduction algorithm (i.e. the LKB model). It follows in the LKB model that uniform EUD irradiation of an OAR results in the same NTCP as the original non-homogeneous distribution. The NTCP equation is therefore represented as a function of EUD. The inverse equation expresses EUD as a function of NTCP and is used to generate a table of EUD versus normal tissue complication probability for the Emami-Burman parameter fits as well as for OAR parameter sets from more recent data.

  17. A new formula for normal tissue complication probability (NTCP) as a function of equivalent uniform dose (EUD)

    International Nuclear Information System (INIS)

    Luxton, Gary; Keall, Paul J; King, Christopher R

    2008-01-01

    To facilitate the use of biological outcome modeling for treatment planning, an exponential function is introduced as a simpler equivalent to the Lyman formula for calculating normal tissue complication probability (NTCP). The single parameter of the exponential function is chosen to reproduce the Lyman calculation to within ∼0.3%, and thus enable easy conversion of data contained in empirical fits of Lyman parameters for organs at risk (OARs). Organ parameters for the new formula are given in terms of Lyman model m and TD 50 , and conversely m and TD 50 are expressed in terms of the parameters of the new equation. The role of the Lyman volume-effect parameter n is unchanged from its role in the Lyman model. For a non-homogeneously irradiated OAR, an equation relates d ref , n, v eff and the Niemierko equivalent uniform dose (EUD), where d ref and v eff are the reference dose and effective fractional volume of the Kutcher-Burman reduction algorithm (i.e. the LKB model). It follows in the LKB model that uniform EUD irradiation of an OAR results in the same NTCP as the original non-homogeneous distribution. The NTCP equation is therefore represented as a function of EUD. The inverse equation expresses EUD as a function of NTCP and is used to generate a table of EUD versus normal tissue complication probability for the Emami-Burman parameter fits as well as for OAR parameter sets from more recent data

  18. Morphological differences in the response of mouse small intestine to radiobiologically equivalent doses of X and neutron irradiation

    International Nuclear Information System (INIS)

    Carr, K.E.; Hamlet, R.; Nias, A.H.; Watt, C.

    1984-01-01

    A scale has been developed to describe the effects of radiation on small intestinal villi. The scale has been used to compare the damage done to the villi in the period 0-5 days after irradiation by X-irradiation or neutron irradiation, using 10 Gy X-rays and 5 Gy neutrons, doses which are radiobiologically equivalent when assessed by the microcolony assay method. Use of the scale indicates that the damage done to the villi by neutrons is greater than that produced by X-rays. This has implications for the interpretation of radiobiological equivalent doses (R.B.E.). Resin light microscopy and transmission electron microscopy (T.E.M.) have also been used to examine small intestinal damage after 10 Gy X-irradiation and 5 Gy neutron irradiation. Differences include variations in crypt shape, mitotic activity and the proportion of crypts which are heavily parasitised. As well as the differences in villous shape which have been reflected in the different values on the scoring system, there are also variations in the response of the constituent cells of the epithelial compartment of the villi. In general, the effect of the neutron irradiation is more severe than that of the X-rays, particularly as would be suggested by a simple quantitation of crypt regeneration

  19. Role of the parameters involved in the plan optimization based on the generalized equivalent uniform dose and radiobiological implications

    International Nuclear Information System (INIS)

    Widesott, L; Strigari, L; Pressello, M C; Landoni, V; Benassi, M

    2008-01-01

    We investigated the role and the weight of the parameters involved in the intensity modulated radiation therapy (IMRT) optimization based on the generalized equivalent uniform dose (gEUD) method, for prostate and head-and-neck plans. We systematically varied the parameters (gEUD max and weight) involved in the gEUD-based optimization of rectal wall and parotid glands. We found that the proper value of weight factor, still guaranteeing planning treatment volumes coverage, produced similar organs at risks dose-volume (DV) histograms for different gEUD max with fixed a = 1. Most of all, we formulated a simple relation that links the reference gEUD max and the associated weight factor. As secondary objective, we evaluated plans obtained with the gEUD-based optimization and ones based on DV criteria, using the normal tissue complication probability (NTCP) models. gEUD criteria seemed to improve sparing of rectum and parotid glands with respect to DV-based optimization: the mean dose, the V 40 and V 50 values to the rectal wall were decreased of about 10%, the mean dose to parotids decreased of about 20-30%. But more than the OARs sparing, we underlined the halving of the OARs optimization time with the implementation of the gEUD-based cost function. Using NTCP models we enhanced differences between the two optimization criteria for parotid glands, but no for rectum wall

  20. Effective dose equivalent to breast fed infants due to artificial and natural radioactivity

    International Nuclear Information System (INIS)

    Campos Venuti, G.; Felici, F.; Grisanti, A.; Grisanti, G.; Risica, S.; Simula, S.

    1988-01-01

    After the Chernobyl accident, the Italian National Institute of Health (ISS), with the collaboration of the Epidemiological Unit of Latium District, started a research devoted to the study of the radioactive contamination of human milk in Latium. The motivation is three-fold, namely: to measure the contamination of a food product utilised for infants in order to assess the corresponding dose to breast-fed infants; to clarify any possible correlations between radioactive levels in human milk and those in the total body; to correlate the levels in human milk with those in the maternal diet

  1. Estimation of equivalent dose and its uncertainty in the OSL SAR protocol when count numbers do not follow a Poisson distribution

    International Nuclear Information System (INIS)

    Bluszcz, Andrzej; Adamiec, Grzegorz; Heer, Aleksandra J.

    2015-01-01

    The current work focuses on the estimation of equivalent dose and its uncertainty using the single aliquot regenerative protocol in optically stimulated luminescence measurements. The authors show that the count numbers recorded with the use of photomultiplier tubes are well described by negative binomial distributions, different ones for background counts and photon induced counts. This fact is then exploited in pseudo-random count number generation and simulations of D e determination assuming a saturating exponential growth. A least squares fitting procedure is applied using dif