  1. Palliative interstitial HDR brachytherapy for recurrent rectal cancer. Implantation techniques and results

    International Nuclear Information System (INIS)

    Kolotas, C.; Roeddiger, S.; Martin, T.; Tselis, N.; Baltas, D.; Zamboglou, N.; Strassmann, G.; Aebersold, D.M.

    2003-01-01

    Purpose: To report the methods and clinical results of CT-based interstitial high-dose-rate (HDR) brachytherapy procedures for the palliative treatment of recurrent rectal cancer. Patients and Methods: A total of 44 brachytherapy implants were performed in 38 patients. CT-guided catheter implants were performed in 34 patients under local anesthesia and sedation, and four patients were implanted intraoperatively. Of 40 CT-guided implants, 20 were done using metallic needles introduced via the sacrum and 20 were transperineal implants of plastic tubes in the presacral region. Postimplant CT scans were used for three-dimensional (3-D) conformal brachytherapy planning. Patients implanted with metallic needles were given a single fraction of 10-15 Gy using HDR 192 Ir, and those who received transperineal implants of plastic catheters were given fractionated brachytherapy, 5 Gy twice daily to a total dose of 30-40 Gy. The median tumor volume was 225 cm 3 with a range of 41-2,103 cm 3 . Results: After a median follow-up of 23.4 months, a total of 13/38 patients were alive. The median postbrachytherapy survival was 15 months with 18 of the 25 deaths due to distant metastases. Tumor response was as follows: 6/38 partial remission, 28/38 stable disease, and 4/38 local progression. A planning target volume (PTV) coverage > 85% was achieved in 42/44 implants. The treatment was well tolerated, and no acute complications were observed. One patient developed a fistula after 8 months. Pain relief was recorded in 34 patients (89.5%), and the median duration of this palliative effect was 5 months with a range of 1-13 months. Conclusions: Interstitial HDR brachytherapy is a valuable tool for the delivery of high doses and achieves effective palliation in recurrent rectal carcinoma. (orig.)

  2. Palliative interstitial HDR brachytherapy for recurrent rectal cancer. Implantation techniques and results

    Energy Technology Data Exchange (ETDEWEB)

    Kolotas, C. [Dept. of Radiation Oncology, Offenbach Hospital, Offenbach (Germany); Dept. of Radio-Oncology, Univ. of Bern, Inselspital, Bern (Switzerland); Roeddiger, S.; Martin, T.; Tselis, N.; Baltas, D.; Zamboglou, N. [Dept. of Radiation Oncology, Offenbach Hospital, Offenbach (Germany); Strassmann, G. [Dept. of Radiotherapy, Univ. Hospital, Philipps Univ., Marburg (Germany); Aebersold, D.M. [Dept. of Radio-Oncology, Univ. of Bern, Inselspital, Bern (Switzerland)

    2003-07-01

    Purpose: To report the methods and clinical results of CT-based interstitial high-dose-rate (HDR) brachytherapy procedures for the palliative treatment of recurrent rectal cancer. Patients and Methods: A total of 44 brachytherapy implants were performed in 38 patients. CT-guided catheter implants were performed in 34 patients under local anesthesia and sedation, and four patients were implanted intraoperatively. Of 40 CT-guided implants, 20 were done using metallic needles introduced via the sacrum and 20 were transperineal implants of plastic tubes in the presacral region. Postimplant CT scans were used for three-dimensional (3-D) conformal brachytherapy planning. Patients implanted with metallic needles were given a single fraction of 10-15 Gy using HDR {sup 192}Ir, and those who received transperineal implants of plastic catheters were given fractionated brachytherapy, 5 Gy twice daily to a total dose of 30-40 Gy. The median tumor volume was 225 cm{sup 3} with a range of 41-2,103 cm{sup 3}. Results: After a median follow-up of 23.4 months, a total of 13/38 patients were alive. The median postbrachytherapy survival was 15 months with 18 of the 25 deaths due to distant metastases. Tumor response was as follows: 6/38 partial remission, 28/38 stable disease, and 4/38 local progression. A planning target volume (PTV) coverage > 85% was achieved in 42/44 implants. The treatment was well tolerated, and no acute complications were observed. One patient developed a fistula after 8 months. Pain relief was recorded in 34 patients (89.5%), and the median duration of this palliative effect was 5 months with a range of 1-13 months. Conclusions: Interstitial HDR brachytherapy is a valuable tool for the delivery of high doses and achieves effective palliation in recurrent rectal carcinoma. (orig.)

  3. Comparison between Japanese and French interstitial brachytherapy for head and neck cancer

    International Nuclear Information System (INIS)

    Nose, Takayuki; Koizumi, Masahiko; Nishiyama, Kinji; Inoue, Toshihiko

    2001-01-01

    Interstitial brachytherapy is the optimal radiotherapy modality for head and neck cancer because the highest dose conformity can be achieved, and implanted tubes can move synchronously with the tumor movement. Compared with radical surgery, interstitial brachytherapy can achieve equivalent local control with less morbidity and less functional deficit. In Japan, because of technical limitations, interstitial brachytherapy has been confined to treatment of small tongue cancers. To improve our head and neck cancer treatment, technical limitations should be eliminated and a wider indication for interstitial brachytherapy should be achieved. In France, interstitial brachytherapy has been technically more developed and widely indicated than in Japan. We analyzed the differences between Japanese (Osaka) and French (Lyon and Nancy) techniques, to improve our interstitial brachytherapy. Implant devices and techniques: French applicators (Longcip 1) are more flexible and more suitable for loop techniques of the soft palate, the base of the tongue, and the vallecula, than applicators available in Japan. Various implant techniques are established especially for the oropharynx in France. Mandibular protection: Lead blocks used in France can more effectively shield the mandible than our silicone spacers. We showed the dosimetric results in an experimental treatment setting. Dose specification: The five-mm dose specification method used in Japan can work only for easy cases, such as small oral tongue cancers and mouth floor cancers. For complicated implants, such as for the oropharynx, the CTV-based dose specification method used in France is essential for sufficient irradiation. Indications: The indication for head and neck interstitial brachytherapy in Japan is limited mostly to small oral tongue cancers. The indication in France is wider, including the oral cavity, the oropharynx, and postoperative cases. We can refine our head and neck cancer treatment if we combine French

  4. A comparison in cosmetic outcome between per-operative interstitial breast implants and delayed interstitial breast implants after external beam radiotherapy

    NARCIS (Netherlands)

    Pieters, Bradley R.; Hart, Augustinus A. M.; Russell, Nicola S.; Jansen, Edwin P. M.; Peterse, Johannes L.; Borger, Jacques; Rutgers, Emiel J. Th

    2003-01-01

    Background and purpose: Interstitial implants for brachytherapy boost in the breast conserving therapy of breast cancer can be performed in two ways; implants during the tumor excision (per-operative implants) or after the external beam therapy (delayed interstitial implants). Differences in

  5. Occupational exposure of professionals during interstitial permanent prostate brachytherapy implants

    International Nuclear Information System (INIS)

    Pirraco, R.; Pereira, A.; Viterbo, T.; Cavaco, A.

    2006-01-01

    Full text of publication follows: Introduction: In this study we present dose measurements for professionals exposed during interstitial 125 I permanent prostate brachytherapy implants. Methods and Materials: The implant technique used was intra operative real time using strand and loose seeds. The professionals inside the operating room are an oncologist, a radiologist, a physicist, a nurse and an anesthesiologist. The oncologist and the physicist contact directly the loaded needle with radioactive seeds and two types of measurements were taken: total body and extremities (finger) dose. The rest of the team operates at long distances, but measurements were made. To measure total body equivalent dose we use a Berthold Umo LB 123 coupled with a LB 1236-H10 detector, and we recorded dose, time and distance from implant location. Finger dosemeters are thermo -luminescent dosimeter (TLD) rings that were controlled over one month. Results: 50 cases (average number of applications per year) were analysed for extremities measurements and 9 cases for total body measurements (in this case, the results were extrapolated for 50 cases), with an average of 26.1 mCi total activity per implant (in a range of 17.4 - 40.3 mCi). The finger dose was 1.8 mSv for the oncologist and 1.9 mSv for the physicist. The interpolation of total body equivalent dose for the oncologist was 24 mSv, for the radiologist 6 mSv and 9 mSv for the physicist. The rest of the team did not receive anything but background radiation. The annual national limit dose for workers is 20 mSv for total body irradiation, and 500 mSv for extremities. Conclusion: In conclusion we may say that during interstitial permanent prostate brachytherapy implants, total doses received for all groups are not significant when compared to annual limits for Portuguese laws 1. Even so, our main goal is always to get the less possible dose (ALARA principle). References: 1. Decreto Lei n. 180/2002 de 8 de Agosto. (authors)

  6. Occupational exposure of professionals during interstitial permanent prostate brachytherapy implants

    Energy Technology Data Exchange (ETDEWEB)

    Pirraco, R.; Pereira, A.; Viterbo, T.; Cavaco, A. [Instituto Portugues de Oncologia Francisco Gentil, Centro R egional de Oncologia do Porto, SA, Porto (Portugal)

    2006-07-01

    Full text of publication follows: Introduction: In this study we present dose measurements for professionals exposed during interstitial 125 I permanent prostate brachytherapy implants. Methods and Materials: The implant technique used was intra operative real time using strand and loose seeds. The professionals inside the operating room are an oncologist, a radiologist, a physicist, a nurse and an anesthesiologist. The oncologist and the physicist contact directly the loaded needle with radioactive seeds and two types of measurements were taken: total body and extremities (finger) dose. The rest of the team operates at long distances, but measurements were made. To measure total body equivalent dose we use a Berthold Umo LB 123 coupled with a LB 1236-H10 detector, and we recorded dose, time and distance from implant location. Finger dosemeters are thermo -luminescent dosimeter (TLD) rings that were controlled over one month. Results: 50 cases (average number of applications per year) were analysed for extremities measurements and 9 cases for total body measurements (in this case, the results were extrapolated for 50 cases), with an average of 26.1 mCi total activity per implant (in a range of 17.4 - 40.3 mCi). The finger dose was 1.8 mSv for the oncologist and 1.9 mSv for the physicist. The interpolation of total body equivalent dose for the oncologist was 24 mSv, for the radiologist 6 mSv and 9 mSv for the physicist. The rest of the team did not receive anything but background radiation. The annual national limit dose for workers is 20 mSv for total body irradiation, and 500 mSv for extremities. Conclusion: In conclusion we may say that during interstitial permanent prostate brachytherapy implants, total doses received for all groups are not significant when compared to annual limits for Portuguese laws 1. Even so, our main goal is always to get the less possible dose (ALARA principle). References: 1. Decreto Lei n. 180/2002 de 8 de Agosto. (authors)

  7. Interstitial prostate brachytherapy. LDR-PDR-HDR

    International Nuclear Information System (INIS)

    Kovacs, Gyoergy; Hoskin, Peter

    2013-01-01

    The first comprehensive overview of interstitial brachytherapy for the management of local or locally advanced prostate cancer. Written by an interdisciplinary team who have been responsible for the successful GEC-ESTRO/EAU Teaching Course. Discusses in detail patient selection, the results of different methods, the role of imaging, and medical physics issues. Prostate brachytherapy has been the subject of heated debate among surgeons and the proponents of the various brachytherapy methods. This very first interdisciplinary book on the subject provides a comprehensive overview of innovations in low dose rate (LDR), high dose rate (HDR), and pulsed dose rate (PDR) interstitial brachytherapy for the management of local or locally advanced prostate cancer. In addition to detailed chapters on patient selection and the use of imaging in diagnostics, treatment guidance, and implantation control, background chapters are included on related medical physics issues such as treatment planning and quality assurance. The results obtained with the different treatment options and the difficult task of salvage treatment are fully discussed. All chapters have been written by internationally recognized experts in their fields who for more than a decade have formed the teaching staff responsible for the successful GEC-ESTRO/EAU Prostate Brachytherapy Teaching Course. This book will be invaluable in informing residents and others of the scientific background and potential of modern prostate brachytherapy. It will also prove a useful source of up-to-date information for those who specialize in prostate brachytherapy or intend to start an interstitial brachytherapy service.

  8. Preliminary results of interstitial [sup 192]Ir brachytherapy for malignant gliomas

    Energy Technology Data Exchange (ETDEWEB)

    Matsumoto, Kengo; Nakagawa, Minoru; Higashi, Hisato [Okayama Univ. (Japan). School of Medicine; and others

    1992-09-01

    Twenty-six patients with recurrent or unremovable malignant gliomas were treated by interstitial brachytherapy with iridium-192 seeds. Stereotactic implantation of the afterloading catheters using the Brown-Roberts-Wells computed tomography (CT)-guided stereotactic system was performed in 24 patients and surgical CT, magnetic resonance imaging, and clinical examination. Tumor regression was seen in 17 patients 1-3 months after implantation. Tumor progression was seen in only three patients. After interstitial brachytherapy, the most commonly observed CT finding was central low density. Median survival time was 18 months after implantation. Autopsies in five patients revealed the delayed effects of radiation injury such as typical vascular changes, microcalcification, and coagulative necrosis in the implant area and tumor recurrence at the periphery. The results suggest that brachytherapy is not curative but prolonged the median survival time by 6 months. (author).

  9. High-dose-rate interstitial brachytherapy for the treatment of penile carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Petera, J.; Odrazka, K.; Zouhar, M.; Bedrosova, J.; Dolezel, M. [Dept. of Oncology and Radiotherapy, Charles Univ. Medical School and Teaching Hospital, Hradec Kralove (Czech Republic)

    2004-02-01

    Background: interstitial low-dose-rate (LDR) brachytherapy allows conservative treatment of T1-T2 penile carcinoma. High-dose-rate (HDR) is often considered to be dangerous for interstitial implants because of a higher risk of complications, but numerous reports suggest that results may be comparable to LDR. Nevertheless, there are no data in the literature available regarding HDR interstitial brachytherapy for carcinoma of the penis. Case report: a 64-year-old man with T1 NO MO epidermoid carcinoma of the glans is reported. Interstitial HDR brachytherapy was performed using the stainless hollow needle technique and a breast template for fixation and good geometry. The dose delivered was 18 x 3 Gy twice daily. Results: after 232 days from brachytherapy, the patient was without any evidence of the tumor, experienced no serious radiation-induced complications, and had a fully functional organ. Conclusion: HDR interstitial brachytherapy is feasible in selected case of penis carcinoma, when careful planning and small single fractions are used. (orig.)

  10. Radiation protection after interstitial permanent prostate brachytherapy implants

    Energy Technology Data Exchange (ETDEWEB)

    Pirraco, R.; Pereira, A.; Cavaco, A. [Instituto Portugues de Oncologia Francisco Gentil - Centro R egional de Oncologia do Porto, SA, Porto (Portugal)

    2006-07-01

    Full text of publication follows: In this study we measure patients radiation exposure dose after interstitial {sup 125}I permanent prostate Brachytherapy implants, and correlate it with dose limits for public, total activity implanted, patient preoperative weight(1), distance between prostate walls and anterior skin surface. Methods and Material: We analyse 20 patients who were implanted with {sup 125}I seeds. The instrument used to measure radiation is a calibrated Berthold Umo LB 123 aco-plated to a LB 1236-H10 detector. Three measurements were taken: at the perineal and anterior pelvic zones on contact with the skin and at 1 m from the patient. The maximum value was taken for all measurements. The dose at a distance of one meter is obtained at anterior pelvic zone, perpendicular to the skin, according to the recommendations of A.A.P.M.(1). The distance between prostate walls was determined using post -operative CT images. Results: The doses at the perineal zone have determined an average of 186 {mu}Sv/h (range: 110 340 {mu}Sv/h) and at surface pelvic zone of 41 {mu}Sv/h (range: 15 103 {mu}Sv/h). The dose at a distance of 1 meter has an average value of 0.4 {mu}Sv/h (range: 0.2 1.0 {mu}Sv/h). The average total activity implanted was 25 mCi (range: 17 38 mCi). The distance between prostate walls and skin pelvic surface of the patients has an average value of 8.9 cm (range: 6.6 -11.5 cm). At a distance of 1 meter from the pelvic zone the dose measured is very low and below dose limits imposed by the European Directive EURATOM 2 and the Portuguese law. For general public to reach annual dose limit (EURATOM - 1 mSv/year) when contacting the pelvic zone, we extrapolate that 4 days (range: 1.6 11.1 days) would be needed, assuming a daily contact period of 6 hours. Conclusion: We established a correlation between the distance of prostate walls to the skin perineal surface and the total dose, but we find no correlation between measured doses, total activity implanted

  11. A comparison in cosmetic outcome between per-operative interstitial breast implants and delayed interstitial breast implants after external beam radiotherapy

    International Nuclear Information System (INIS)

    Pieters, Bradley R.; Hart, Augustinus A.M.; Russell, Nicola S.; Jansen, Edwin P.M.; Peterse, Johannes L.; Borger, Jacques; Rutgers, Emiel J.Th.

    2003-01-01

    Background and purpose: Interstitial implants for brachytherapy boost in the breast conserving therapy of breast cancer can be performed in two ways; implants during the tumor excision (per-operative implants) or after the external beam therapy (delayed interstitial implants). Differences in cosmetic outcome were investigated. Patients and methods: Cosmetic results in 47 patients having a per-operative implant were compared to 123 patients having a delayed interstitial implant in a matched case-control study. Cosmesis was scored on a four-point-scale varying from 0 (excellent) to 3 (poor). Results: After mean follow-up of 63 months, three observers found no difference in cosmetic outcome between the two groups after adjustment for variables found to be related with cosmesis (difference in mean score 0.50, P=0.26). Implant volume at 100% isodose was not found to differ (P=0.084) between the per-operative group (mean 102 cm 3 , S.D. 34 cm 3 ) and the delayed group (mean 93 cm 3 , S.D. 29 cm 3 ). Conclusions: Performing per-operative implants has not led to smaller implants. The method of performing brachytherapy does not result in marked differences in cosmetic outcome

  12. Implant strategies for endocervical and interstitial ultrasound hyperthermia adjunct to HDR brachytherapy for the treatment of cervical cancer

    International Nuclear Information System (INIS)

    Wootton, Jeffery H; Prakash, Punit; Hsu, I-Chow Joe; Diederich, Chris J

    2011-01-01

    Catheter-based ultrasound devices provide a method to deliver 3D conformable heating integrated with HDR brachytherapy delivery. Theoretical characterization of heating patterns was performed to identify implant strategies for these devices which can best be used to apply hyperthermia to cervical cancer. A constrained optimization-based hyperthermia treatment planning platform was used for the analysis. The proportion of tissue ≥41 deg. C in a hyperthermia treatment volume was maximized with constraints T max ≤ 47 deg. C, T rectum ≤ 41.5 deg. C, and T bladder ≤ 42.5 deg. C. Hyperthermia treatment was modeled for generalized implant configurations and complex configurations from a database of patients (n = 14) treated with HDR brachytherapy. Various combinations of endocervical (360 0 or 2 x 180 0 output; 6 mm OD) and interstitial (180 0 , 270 0 , or 360 0 output; 2.4 mm OD) applicators within catheter locations from brachytherapy implants were modeled, with perfusion constant (1 or 3 kg m -3 s -1 ) or varying with location or temperature. Device positioning, sectoring, active length and aiming were empirically optimized to maximize thermal coverage. Conformable heating of appreciable volumes (>200 cm 3 ) is possible using multiple sectored interstitial and endocervical ultrasound devices. The endocervical device can heat >41 deg. C to 4.6 cm diameter compared to 3.6 cm for the interstitial. Sectored applicators afford tight control of heating that is robust to perfusion changes in most regularly spaced configurations. T 90 in example patient cases was 40.5-42.7 deg. C (1.9-39.6 EM 43deg.C ) at 1 kg m -3 s -1 with 10/14 patients ≥41 deg. C. Guidelines are presented for positioning of implant catheters during the initial surgery, selection of ultrasound applicator configurations, and tailored power schemes for achieving T 90 ≥ 41 deg. C in clinically practical implant configurations. Catheter-based ultrasound devices, when adhering to the guidelines, show

  13. Dose optimization of intra-operative high dose rate interstitial brachytherapy implants for soft tissue sarcoma

    Directory of Open Access Journals (Sweden)

    Jamema Swamidas

    2009-01-01

    Full Text Available Objective : A three dimensional (3D image-based dosimetric study to quantitatively compare geometric vs. dose-point optimization in combination with graphical optimization for interstitial brachytherapy of soft tissue sarcoma (STS. Materials and Methods : Fifteen consecutive STS patients, treated with intra-operative, interstitial Brachytherapy, were enrolled in this dosimetric study. Treatment plans were generated using dose points situated at the "central plane between the catheters", "between the catheters throughout the implanted volume", at "distances perpendicular to the implant axis" and "on the surface of the target volume" Geometrically optimized plans had dose points defined between the catheters, while dose-point optimized plans had dose points defined at a plane perpendicular to the implant axis and on the target surface. Each plan was graphically optimized and compared using dose volume indices. Results : Target coverage was suboptimal with coverage index (CI = 0.67 when dose points were defined at the central plane while it was superior when the dose points were defined at the target surface (CI=0.93. The coverage of graphically optimized plans (GrO was similar to non-GrO with dose points defined on surface or perpendicular to the implant axis. A similar pattern was noticed with conformity index (0.61 vs. 0.82. GrO were more conformal and less homogeneous compared to non-GrO. Sum index was superior for dose points defined on the surface of the target and relatively inferior for plans with dose points at other locations (1.35 vs. 1.27. Conclusions : Optimization with dose points defined away from the implant plane and on target results in superior target coverage with optimal values of other indices. GrO offer better target coverage for implants with non-uniform geometry and target volume.

  14. Directional interstitial brachytherapy from simulation to application

    Science.gov (United States)

    Lin, Liyong

    Organs at risk (OAR) are sometimes adjacent to or embedded in or overlap with the clinical target volume (CTV) to be treated. The purpose of this PhD study is to develop directionally low energy gamma-emitting interstitial brachytherapy sources. These sources can be applied between OAR to selectively reduce hot spots in the OARs and normal tissues. The reduction of dose over undesired regions can expand patient eligibility or reduce toxicities for the treatment by conventional interstitial brachytherapy. This study covers the development of a directional source from design optimization to construction of the first prototype source. The Monte Carlo code MCNP was used to simulate the radiation transport for the designs of directional sources. We have made a special construction kit to assemble radioactive and gold-shield components precisely into D-shaped titanium containers of the first directional source. Directional sources have a similar dose distribution as conventional sources on the treated side but greatly reduced dose on the shielded side, with a sharp dose gradient between them. A three-dimensional dose deposition kernel for the 125I directional source has been calculated. Treatment plans can use both directional and conventional 125I sources at the same source strength for low-dose-rate (LDR) implants to optimize the dose distributions. For prostate tumors, directional 125I LDR brachytherapy can potentially reduce genitourinary and gastrointestinal toxicities and improve potency preservation for low risk patients. The combination of better dose distribution of directional implants and better therapeutic ratio between tumor response and late reactions enables a novel temporary LDR treatment, as opposed to permanent or high-dose-rate (HDR) brachytherapy for the intermediate risk T2b and high risk T2c tumors. Supplemental external-beam treatments can be shortened with a better brachytherapy boost for T3 tumors. In conclusion, we have successfully finished the

  15. Interstitial brachytherapy in carcinoma of the penis

    Energy Technology Data Exchange (ETDEWEB)

    Chaudhary, A.J.; Ghosh, S.; Bhalavat, R.L. [Tata Memorial Hospital, Mumbai (India). Dept. of Radiation Oncology; Kulkarni, J.N. [Tata Memorial Hospital, Mumbai (India). Dept. of Surgery; Sequeira, B.V.E. [Tata Memorial Hospital, Mumbai (India). Dept. of Medical Physics

    1999-01-01

    Aim: Keeping in line with the increasing emphasis on organ preservation, we at the Tata Memorial Hospital have evaluated the role of Ir-192 interstitial implant as regards local control, functional and cosmetic outcome in early as well as locally recurrent carcinoma of the distal penis. Patients and Methods: From October 1988 to December 1996, 23 patients with histopathologically proven cancer of the penis were treated with radical radiation therapy using Ir-192 temporary interstitial implant. Our patients were in the age group of 20 to 60 years. The primary lesions were T1 and 7, T2 in 7 and recurrent in 9 patients. Only 7 patients had palpable groin nodes at presentation, all of which were pathologically negative. The median dose of implant was 50 Gy (range 40 to 60 Gy), using the LDR afterloading system and the Paris system of implant rules for dosimetry. Follow-up ranged from 4 to 117 months (median 24 months). Results: At last follow-up 18 of the 23 patients remained locally controlled with implant alone. Three patients failed only locally, 2 locoregionally and 1 only at the groin. Of the 5 patients who failed locally, 4 were successfully salvaged with partial penectomy and remained controlled when last seen. Local control with implant alone at 8 years was 70% by life table analysis. The patients had excellent functional and cosmetic outcome. We did not record any case of skin or softtissue necrosis. Only 2 patients developed meatal stenosis, both of which were treated endoscopically. Conclusion: Our results lead us to interpret that interstitial brachytherapy with Ir-192 offers excellent local control rates with preservation of organ and function. Penectomy can be reserved as a means for effective salvage. (orig.) [Deutsch] Ziel: Das Prinzip des Organerhalts gewinnt in der Onkologie zunehmend an Bedeutung. Ziel dieser Untersuchung war es, die Rolle der interstitiellen Brachytherapie mit Ir-192 zur Behandlung des fruehen und rezidivierten Peniskarzinoms zu

  16. Interstitial brachytherapy with 192-IR in treatment of recurrent malignant primary brain tumors

    International Nuclear Information System (INIS)

    Cardenes, R.; Martinez, R.; Victoria, C.; Nunez, L.; Clavo, B.; Sancedo, G.

    1994-01-01

    Seven patients with recurrent malignant primary brain tumors after surgery and radiation therapy were treated at the Clinica Puerta de Hierro (Madrid) by interstitial brachytherapy with 192-Ir sources. Implantations were performed using computerized tomography and dose prescription were determined following the Paris system rules for interstitial implants. The means dose deliberated was 50 to 65 Gy to the reference isodoses. At the last follow-up all patients except for one are alive and without evidence of progression of the disease. (Author) 35 refs

  17. Clinical Investigations of a CT-based reconstruction and 3D-Treatment planning system in interstitial brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Kolotas, C; Zamboglou, N [Strahlenklinik, Stadtische Kliniken Offenbach, Offenbach (Germany)

    1999-12-31

    Purpose: Development, application and evaluation of a CT-guided implantation technique and a fully CT based treatment planning procedure for brachytherapy. Methods and Materials : A brachytherapy procedure based on CT-guided implantation technique and CT based treatment planning has been developed and clinically evaluated. For this purpose a software system (PROMETHEUS) for the 3D reconstruction of brachytherapy catheters and patient anatomy using only CT scans has been developed. An interface for the Nucletron Plato BPS treatment planning system for the optimisation and calculation of dose distribution has been devised. The planning target volume(s) are defined as sets of points using contouring tools and are for optimisation of the 3D dose distribution. Dose-volume histogram-based analysis of the dose distribution enables a clinically realistic evaluation of the brachytherapy application to be made. The CT-guided implantation of catheters and the CT-based treatment planning procedure has been performed for interstitial brachytherapy and for different tumour and anatomical localizations in 197 patients between 1996 and 1997. Results : The accuracy of the CT reconstruction was tested using a quality assurance phantom an an interstitial implant of 12 needles and compared with the results of reconstruction using radiographs[hs. Both methods give comparable results with regard to accuracy. The CT based reconstruction was faster. Clinical feasibility has been proven in pre-irradiated recurrences of brain tumour, in pre-treated recurrences or metastatic disease, and in breast carcinomas. The tumour volume treated ranged from 5.1 - 2741 cm3. Analysis of the implant quality showed a slight significant lower COIN value for the bone implants, but no differences in respect to the planning target volume. Conclusions : With the integration of CT imaging in the treatment planning and documentation of brachytherapy, we have a new CT based quality assurance method to evaluate

  18. Clinical Investigations of a CT-based reconstruction and 3D-Treatment planning system in interstitial brachytherapy

    International Nuclear Information System (INIS)

    Kolotas, C.; Zamboglou, N.

    1998-01-01

    Purpose: Development, application and evaluation of a CT-guided implantation technique and a fully CT based treatment planning procedure for brachytherapy. Methods and Materials : A brachytherapy procedure based on CT-guided implantation technique and CT based treatment planning has been developed and clinically evaluated. For this purpose a software system (PROMETHEUS) for the 3D reconstruction of brachytherapy catheters and patient anatomy using only CT scans has been developed. An interface for the Nucletron Plato BPS treatment planning system for the optimisation and calculation of dose distribution has been devised. The planning target volume(s) are defined as sets of points using contouring tools and are for optimisation of the 3D dose distribution. Dose-volume histogram-based analysis of the dose distribution enables a clinically realistic evaluation of the brachytherapy application to be made. The CT-guided implantation of catheters and the CT-based treatment planning procedure has been performed for interstitial brachytherapy and for different tumour and anatomical localizations in 197 patients between 1996 and 1997. Results : The accuracy of the CT reconstruction was tested using a quality assurance phantom an an interstitial implant of 12 needles and compared with the results of reconstruction using radiographs[hs. Both methods give comparable results with regard to accuracy. The CT based reconstruction was faster. Clinical feasibility has been proven in pre-irradiated recurrences of brain tumour, in pre-treated recurrences or metastatic disease, and in breast carcinomas. The tumour volume treated ranged from 5.1 - 2741 cm3. Analysis of the implant quality showed a slight significant lower COIN value for the bone implants, but no differences in respect to the planning target volume. Conclusions : With the integration of CT imaging in the treatment planning and documentation of brachytherapy, we have a new CT based quality assurance method to evaluate

  19. 192Ir high dose rate (HDR) interstitial brain implant: optimisation

    International Nuclear Information System (INIS)

    Tyagi, Anuj; Singh, Dinesh; Chitra, S.; Gupta, J.P.

    2001-01-01

    The new modality of stepping source dosimetry system (SSDs) illustrates a remarkable improvement in attaining the uniform and homogeneous dose distribution within the target volume. The technique enables the physicist to correct for a certain amount of misplacement or curvature of implant geometry. The short course of brachytherapy provides good palliation in terms of functional improvements with low and acceptable toxicity in high-grade glioma. With continual refinements of the technique, brachytherapy performed by a skilled brachytherapy team offers an opportunity to improve patient survival and quality of life. Since 1997, micro selectron HDR 192 Ir treatments are done including gynecological, oesophageal, breast, surface mould, soft tissue sarcoma (STS) and brain in our hospital. In this paper, procedure of interstitial brain implant in glioma as implant technique, simulation and treatment planning will be discussed

  20. An overview of interstitial brachytherapy and hyperthermia

    International Nuclear Information System (INIS)

    Brandt, B.B.; Harney, J.

    1989-01-01

    Interstitial thermoradiotherapy, an experimental cancer treatment that combines interstitial radiation implants (brachytherapy) and interstitial hyperthermia, is in the early stages of investigation. In accordance with the procedure used in a current national trial protocol, a 60-minute hyperthermia treatment is administered after catheters are placed into the tumor area while the patient is under general anesthesia. This is immediately followed by loading of radioactive Iridium-192 seeds into the catheters for a defined period of time. Once the prescribed radiation dose is delivered, the radioactive sources are removed and a second, 60-minute hyperthermia treatment is administered. Clinical trials with hyperthermia in combination with radiation have increased in recent years. Nurses caring for these patients need to become more knowledgeable about this investigational therapy. This paper provides an overview of the biologic rationale for this therapy, as well as a description of the delivery method and clinical application. Specific related nursing interventions are defined in a nursing protocol.23 references

  1. Evaluation of failure modes of computerized planning phase of interstitial implants with high dose rate brachytherapy using HFMEA

    International Nuclear Information System (INIS)

    Biazotto, Bruna; Tokarski, Marcio

    2014-01-01

    This paper evaluates the failure modes of the computerized planning step in interstitial implants with high dose rate brachytherapy. The prospective tool of risk management Health Care Failure Mode and Effects Analysis (HFMEA) was used. Twelve subprocesses were identified, and 33 failure modes of which 21 justified new safety actions, and 9 of them were intolerable risks. The method proved itself useful in identifying failure modes, but laborious and subjective in their assessment. The main risks were due to human factors, which require training and commitment of management to their mitigation. (author)

  2. Open magnetic resonance imaging using titanium-zirconium needles: improved accuracy for interstitial brachytherapy implants?

    International Nuclear Information System (INIS)

    Popowski, Youri; Hiltbrand, Emile; Joliat, Dominique; Rouzaud, Michel

    2000-01-01

    Purpose: To evaluate the benefit of using an open magnetic resonance (MR) machine and new MR-compatible needles to improve the accuracy of brachytherapy implants in pelvic tumors. Methods and Materials: The open MR machine, foreseen for interventional procedures, allows direct visualization of the pelvic structures that are to be implanted. For that purpose, we have developed MR- and CT-compatible titanium-zirconium (Ti-Zr) brachytherapy needles that allow implantations to be carried out under the magnetic field. In order to test the technical feasibility of this new approach, stainless steel (SS) and Ti-Zr needles were first compared in a tissue-equivalent phantom. In a second step, two patients implanted with Ti-Zr needles in the brachytherapy operating room were scanned in the open MR machine. In a third phase, four patients were implanted directly under open MR control. Results: The artifacts induced by both materials were significantly different, strongly favoring the Ti-Zr needles. The implantation in both first patients confirmed the excellent quality of the pictures obtained with the needles in vivo and showed suboptimal implant geometry in both patients. In the next 4 patients, the tumor could be punctured with excellent accuracy, and the adjacent structures could be easily avoided. Conclusion: We conclude that open MR using MR-compatible needles is a very promising tool in brachytherapy, especially for pelvic tumors

  3. Implant strategies for endocervical and interstitial ultrasound hyperthermia adjunct to HDR brachytherapy for the treatment of cervical cancer

    Science.gov (United States)

    Wootton, Jeffery H.; Prakash, Punit; Hsu, I.-Chow Joe; Diederich, Chris J.

    2011-07-01

    Catheter-based ultrasound devices provide a method to deliver 3D conformable heating integrated with HDR brachytherapy delivery. Theoretical characterization of heating patterns was performed to identify implant strategies for these devices which can best be used to apply hyperthermia to cervical cancer. A constrained optimization-based hyperthermia treatment planning platform was used for the analysis. The proportion of tissue >=41 °C in a hyperthermia treatment volume was maximized with constraints Tmax 200 cm3) is possible using multiple sectored interstitial and endocervical ultrasound devices. The endocervical device can heat >41 °C to 4.6 cm diameter compared to 3.6 cm for the interstitial. Sectored applicators afford tight control of heating that is robust to perfusion changes in most regularly spaced configurations. T90 in example patient cases was 40.5-42.7 °C (1.9-39.6 EM43 °C) at 1 kg m-3 s-1 with 10/14 patients >=41 °C. Guidelines are presented for positioning of implant catheters during the initial surgery, selection of ultrasound applicator configurations, and tailored power schemes for achieving T90 >= 41 °C in clinically practical implant configurations. Catheter-based ultrasound devices, when adhering to the guidelines, show potential to generate conformal therapeutic heating ranging from a single endocervical device targeting small volumes local to the cervix (directional interstitial applicators in the lateral periphery to target much larger volumes (6 cm radial), while preferentially limiting heating of the bladder and rectum.

  4. Low dose rate Ir-192 interstitial brachytherapy for prostate cancer

    Energy Technology Data Exchange (ETDEWEB)

    Oki, Yosuke; Dokiya, Takushi; Yorozu, Atsunori; Suzuki, Takayuki; Saito, Shiro; Monma, Tetsuo; Ohki, Takahiro [National Tokyo Medical Center (Japan); Murai, Masaru; Kubo, Atsushi

    2000-04-01

    From December 1997 through January 1999, fifteen prostatic cancer patients were treated with low dose rate Ir-192 interstitial brachytherapy using TRUS and perineal template guidance without external radiotherapy. Up to now, as no apparent side effects were found, the safety of this treatment is suggested. In the future, in order to treat prostatic cancer patients with interstitial brachytherapy using I-125 or Pd-103, more investigation for this low dose rate Ir-192 interstitial brachytherapy is needed. (author)

  5. Permanent interstitial implantation of 125Iodine seed for thoracic malignant tumors

    International Nuclear Information System (INIS)

    Xu Zhongheng; Qian Yongyue; Wu Jinchang; Liu Zengli

    2002-01-01

    Objective: To observe effect of 125 Iodine sed on interstitial brachytherapy of patient with thoracic malignant tumor. Methods: 125 Iodine seed were inserted into the target tissue and permanent left there for brachytherapy in 6 cases of thoracic malignant tumors, which including lung cancer, Pancoast's tumour, mediastinal malignant schwannoma. Results: All cases were rehabilitated shortly after operation. The implanted lesions remained controlled now and in dead patients. No radiation-related and 125 Iodine seed-related complications occurred. Conclusion: Brachytherapy by implantation of 125 Iodine seeds of remained tumor tissue in patients with thoracic malignant tumor after operation has a satisfactory outcome. This therapy can control local recurrent of thoracic malignant tumor. But the results in long term should be studied further

  6. A study on applying Ra needle interstitial brachytherapy for oral cancer

    International Nuclear Information System (INIS)

    Yoshida, Shunichi; Komiya, Yoshiaki; Uchida, Ikuhiro; Tashiro, Kazuyoshi

    1999-01-01

    To investigate applicability of Ra needle interstitial brachytherapy, 93 cases of oral squamous carcinoma were examined. The patients underwent Ra needle interstitial brachytherapy as thorough therapy in our hospital. The criteria of applying Ra needle interstitial brachytherapy was diameter of within 5 cm and depth within 2 cm of tumor size. Ra needle interstitial brachytherapy was applied to 82 cases of tongue, 10 cases of oral floor and one case of lower lip carcinomas. The local control rate was 92.5%, and secondary neck metastasis was 32.3% in all cases applied Ra needle interstitial brachytherapy. The results were not bad compared with surgical treatment. However, the 5-year cumulative survival rate was 64.9%, which was not good enough at the result to obtain a good local control rate and secondary neck metastasis rate. The result was relative to low treatment result of local recurrence cases with Ra needle interstitial brachytherapy. To improve the result, it is important to distinguish local recurrence from radioinduced ulcer, and to start early secondary treatment. The cases in which cervical lymph node metastasis was found as the first examination underwent neck dessection after Ra needle interstitial brachytherapy. The 5-year cumulative survival rate was 83.3% in N1 cases and 40.6% in N2 cases, and the result of N2 cases was poorer than N1 cases with a significant difference. The results indicate that a needle having a diameter of within 5 cm, depth of within 2 cm and less than N1 can be applied during Ra needle interstitial brachytherapy for complete cure of cancer. (author)

  7. Implementation of 3D-virtual brachytherapy in the management of breast cancer: a description of a new method of interstitial brachytherapy

    International Nuclear Information System (INIS)

    Vicini, Frank A.; Jaffray, David A.; Horwitz, Eric M.; Edmundson, Gregory K.; DeBiose, David A.; Kini, Vijay R.; Martinez, Alvaro A.

    1998-01-01

    Purpose: We present the initial description of a new technique of interstitial breast brachytherapy in which a computer-generated image of an implant template is applied virtually to serial-computed tomography (CT) scan images of a patient's breast. Optimal placement of the virtual template around the CT images of the proposed target volume provides the physician with a preplan for improved positioning of implant needles around the actual target volume intraoperatively. Methods and Materials: Since March of 1993, 110 patients with early-stage breast cancer were entered onto a protocol of low or high dose rate brachytherapy as the sole radiation modality for part of their breast-conserving therapy. To improve the accuracy and reproducibility of target volume coverage in patients with a closed lumpectomy cavity, 11 of these implants were performed using the virtual brachytherapy technique. The virtual implant procedure was performed by first placing radiopaque skin markers on the breast surface for reference on the CT image and ultimately as intraoperative landmarks for the placement of implant needles. A CT scan of the breast was then performed and the target volume outlined on each CT scan slice by the physician. A virtual image of the brachytherapy template was then positioned around the CT image of the target volume to achieve an idealized implant with optimal coverage. The projected entrance and exit points of all needles on the skin of the breast (from the idealized virtual implant) were then identified (by perspective rendering of multiple 3D views) and hard-copy images taken to the operating room. The implant was then constructed by referencing the virtual implant images (needle entrance and exit points) to the radiopaque skin markers on the breast. After the implant was completed, a CT scan of the breast with the template catheters or needles in position was taken for comparison of the actual target volume coverage with the virtual implant generated

  8. Californium-252 interstitial implants in carcinoma of the tongue

    International Nuclear Information System (INIS)

    Vtyurin, B.M.; Ivanov, V.N.; Medvedev, V.S.; Galantseva, G.F.; Abdulkadyrov, S.A.; Ivanova, L.F.; Petrovskaya, G.A.; Plichko, V.I.

    1985-01-01

    A clinical study using 252 Cf sources in brachytherapy of tumors began in the Research Institute of Medical Radiology of the Academy of Medical Sciences of the USSR in 1973. 252 Cf afterloading cells were utilized by the method of simple afterloading. Dosimetry and radiation protection of medical personnel were developed. To substantiate optimal therapeutic doses of 252 Cf neutrons, a correlation of dose, time, and treatment volume factors with clinical results of 252 Cf interstitial implants in carcinoma of the tongue for 47 patients with a minimum follow-up period of 1 year was studied. Forty-nine interstitial implants have been performed. Seventeen patients received 252 Cf implants alone (Group I), 17 other patients received 252 Cf implants in combination with external radiation (Group II), and 15 patients were treated with interstitial implants for recurrent or residual tumors (Groups III). Complete regression of carcinoma of the tongue was obtained in 48 patients (98%). Thirteen patients (27%) developed radiation necrosis. The therapeutic dose of neutron radiation from 252 Cf sources in interstitial radiotherapy of primary tongue carcinomas (Group I) was found to be 7 to 9 Gy. Optimal therapeutic neutron dose in combined interstitial and external radiotherapy of primary tumors (Group II) was 5 to 6 Gy with an external radiation dose of 40 Gy. For recurrent and residual tumors (Group III), favorable results were obtained with tumor doses of 6.5 to 7 Gy

  9. Pelvic interstitial brachytherapy - improving the therapeutic ratio with magnetic resonance imaging and optimization

    International Nuclear Information System (INIS)

    Swift, Patrick S.; Hricak, Hedvig; Forstner, Rosemary; Powell, C. Bethan; Purser, Phil; Weaver, Keith; Phillips, Theodore L.

    1996-01-01

    Introduction Interstitial brachytherapy in the pelvic region is often hampered by the radiation oncologist's inability to precisely differentiate tumor versus normal tissue during the planning and implantation procedures, often resulting in either excessive or incomplete coverage of tumor volume. The marked improvement in pelvic imaging seen with magnetic resonance, in conjunction with isodose optimization programs for remote-afterloading units, has created an opportunity to significantly improve the therapeutic ratio. Methods From 1992-1995, 23 interstitial perineal templates were performed in 22 patients with pelvic malignancies, using the pulsed low-dose-rate Selectron with dose optimization. MR imaging was performed immediately prior to the implant, with a MUPIT placed against the perineum and a vaginal obturator in place. These images were used for tumor volume measurements, determination of the number, depth and angle of needles required for the implant, and identification of position of normal tissues (rectum, small bowel, bladder) relative to the tumor. After implantation of stainless steel needles, orthogonal radiographs were obtained for isodose calculation, and planning carried out with isodose optimization. Patients were followed closely on a routine schedule, until time of last visit or until death. Every effort possible was made to assess local disease status at time of death. Results Sixteen patients with primary disease (14 cervix, 1 vulva, 1 vagina) and 6 with recurrent (2 with prior radiation) were implanted, all but 3 with curative intent. Nine patients with advanced cervix or vulvar cancer received concomitant chemotherapy (5FU + platinum or mitomycin-C) with the external beam therapy. At a median follow-up of 18.1 months for all cases, only three patients have failed locally for an actuarial local control of 85% at 1.5 years. Nine patients are alive and free of disease, 8 are alive with distant disease only (mean follow-up of 19.1 months), 2

  10. Permanent I-125 interstitial implant in the management of high grade CNS malignancies in children

    International Nuclear Information System (INIS)

    Vaishampayan, N.; Zamorano, L.; Aronin, P.; Gaspar, L.; Canady, A.; Lattin, P.; Ezzell, G.; Yakar, D.; Chungbin, S.; Fontanesi, J.

    1996-01-01

    Purpose/Objective: To evaluate the efficacy and complications associated with the use of permanent I-125 interstitial implants in children with high grade CNS malignancies. Materials and Methods: Between May of 1990 and September of 1994, fourteen children received permanent I-125 interstitial implant brachytherapy as initial therapy (n=8) or at time of recurrence (n=6). Histologies included Glioblastoma Multiforme (n=2), Anaplastic Astrocytoma (n=9) and others (n=3). Pre-implant surgical procedures included: Gross Total Resection (n=2), Subtotal Resection (n=8) or Biopsy alone (n=4). Six patients received pre-implant external beam irradiation (dose range 3,500-6500 cGy) and three patients received post-implant external beam irradiation (dose range 5,040-5,060 cGy). Implant dose range was 8,294-10,368 cGy over the lifetime of the implant (median 10,368 cGy). Results: At last follow-up (median 17.5 months; range 4-56 months), eight children were alive. Six out of the eight had no evidence of disease progression while the remaining had radiologic evidence of progression. Implant complications (n=2) included skin necrosis and bone flap infection. Conclusions: Based on this initial review, we continue to investigate the use of permanent I-125 interstitial brachytherapy in the treatment of high grade CNS malignancies in children and will discuss and compare these results with those of other 'Boost' series

  11. Dose optimisation in single plane interstitial brachytherapy

    DEFF Research Database (Denmark)

    Tanderup, Kari; Hellebust, Taran Paulsen; Honoré, Henriette Benedicte

    2006-01-01

    patients,       treated for recurrent rectal and cervical cancer, flexible catheters were       sutured intra-operatively to the tumour bed in areas with compromised       surgical margin. Both non-optimised, geometrically and graphically       optimised CT -based dose plans were made. The overdose index...... on the       regularity of the implant, such that the benefit of optimisation was       larger for irregular implants. OI and HI correlated strongly with target       volume limiting the usability of these parameters for comparison of dose       plans between patients. CONCLUSIONS: Dwell time optimisation significantly......BACKGROUND AND PURPOSE: Brachytherapy dose distributions can be optimised       by modulation of source dwell times. In this study dose optimisation in       single planar interstitial implants was evaluated in order to quantify the       potential benefit in patients. MATERIAL AND METHODS: In 14...

  12. Navigation system for interstitial brachytherapy

    International Nuclear Information System (INIS)

    Strassmann, G.; Kolotas, C.; Heyd, R.

    2000-01-01

    The purpose of the stud was to develop a computed tomography (CT) based electromagnetic navigation system for interstitial brachytherapy. This is especially designed for situations when needles have to be positioned adjacent to or within critical anatomical structures. In such instances interactive 3D visualisation of the needle positions is essential. The material consisted of a Polhemus electromagnetic 3D digitizer, a Pentium 200 MHz laptop and a voice recognition for continuous speech. In addition, we developed an external reference system constructed of Perspex which could be positioned above the tumour region and attached to the patient using a non-invasive fixation method. A specially designed needle holder and patient bed were also developed. Measurements were made on a series of phantoms in order to study the efficacy and accuracy of the navigation system. The mean navigation accuracy of positioning the 20.0 cm length metallic needles within the phantoms was in the range 2.0-4.1 mm with a maximum of 5.4 mm. This is an improvement on the accuracy of a CT-guided technique which was in the range 6.1-11.3 mm with a maximum of 19.4 mm. The mean reconstruction accuracy of the implant geometry was 3.2 mm within a non-ferromagnetic environment. We found that although the needles were metallic this did not have a significant influence. We also found for our experimental setups that the CT table and operation table non-ferromagnetic parts had no significant influence on the navigation accuracy. This navigation system will be a very useful clinical tool for interstitial brachytherapy applications, particularly when critical structures have to be avoided. It also should provide a significant improvement on our existing technique

  13. American Brachytherapy Society consensus report for accelerated partial breast irradiation using interstitial multicatheter brachytherapy.

    Science.gov (United States)

    Hepel, Jaroslaw T; Arthur, Douglas; Shaitelman, Simona; Polgár, Csaba; Todor, Dorin; Zoberi, Imran; Kamrava, Mitchell; Major, Tibor; Yashar, Catheryn; Wazer, David E

    To develop a consensus report for the quality practice of accelerated partial breast irradiation (APBI) using interstitial multicatheter brachytherapy (IMB). The American Brachytherapy Society Board appointed an expert panel with clinical and research experience with breast brachytherapy to provide guidance for the current practice of IMB. This report is based on a comprehensive literature review with emphasis on randomized data and expertise of the panel. Randomized trials have demonstrated equivalent efficacy of APBI using IMB compared with whole breast irradiation for select patients with early-stage breast cancer. Several techniques for placement of interstitial catheters are described, and importance of three-dimensional planning with appropriate optimization is reviewed. Optimal target definition is outlined. Commonly used dosing schemas include 50 Gy delivered in pulses of 0.6-0.8 Gy/h using pulsed-dose-rate technique and 34 Gy in 10 fractions, 32 Gy in eight fractions, or 30 Gy in seven fractions using high-dose-rate technique. Potential toxicities and strategies for toxicity avoidance are described in detail. Dosimetric constraints include limiting whole breast volume that receives ≥50% of prescription dose to 0.75 (>0.85 preferred), V 150  < 45 cc, and V 200  < 14 cc. Using an optimal implant technique coupled with optimal planning and appropriate dose constraints, a low rate of toxicity and a good-to-excellent cosmetic outcome of ≥90% is expected. IMB is an effective technique to deliver APBI for appropriately selected women with early-stage breast cancer. This consensus report has been created to assist clinicians in the appropriate practice of APBI using IMB. Copyright © 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  14. Factors affecting radiation injury after interstitial brachytherapy for brain tumors

    International Nuclear Information System (INIS)

    Leibel, S.A.; Gutin, P.H.; Davis, R.L.

    1991-01-01

    The effects of brachytherapy on normal brain tissue are not easily delineated in the clinical setting because of the presence of concurrent radiation-induced changes in the coexistent brain tumor. Sequential morphologic studies performed after the implantation of radioactive sources into the brains of experimental animals have provided a better understanding of the character and magnitude of the structural changes produced by interstitial irradiation on normal brain tissue. Furthermore, the clinical experience accumulated thus far provides not only relevant information, but also some guidelines for future treatment policies. In this paper, the authors summarize the experimental findings and review the pathologic and clinical features of brain injury caused by interstitial brachytherapy. A number of studies in the older literature examined the effects of radioisotopes such as radium-226 (38--43), radon-22 (44--46), gold-198 (29,47--50), tantalum-182 (29,51,52) yttrium-9- (50,53,54), and cobalt-60 (29,50,55). This review is restricted to low- and high-activity encapsulated iodine-125 ( 125 I) and iridium-192 ( 192 Ir), the isotopes that are most commonly used in current clinical practice

  15. 3D-CT implanted interstitial brachytherapy for T2b nasopharyngeal carcinoma

    International Nuclear Information System (INIS)

    Ren, Yu-Feng; Gao, Yuan-Hong; Cao, Xin-Ping; Ye, Wei-Jun; Teh, Bin S

    2010-01-01

    To compare the results of external beam radiotherapy in combination with 3D- computed tomography (CT)-implanted interstitial high dose rate brachytherapy (ERT/3D-HDR-BT) versus conventional external beam radiotherapy (ERT) for the treatment of stage T2b nasopharyngeal carcinoma (NPC). Forty NPC patients diagnosed with stage T2b NPC were treated with ERT/3D-HDR-BT under local anesthesia. These patients received a mean dose of 60 Gy, followed by 12-20 Gy administered by 3D-HDR-BT. Another 101 patients diagnosed with non-metastatic T2b NPC received a mean dose of 68 Gy by ERT alone during the same period. Patients treated with ERT/3D-HDR-BT versus ERT alone exhibited an improvement in their 5-y local failure-free survival rate (97.5% vs. 80.2%, P = 0.012) and disease-free survival rate (92.5% vs. 73.3%, P = 0.014). Using multivariate analysis, administration of 3D-HDR-BT was found to be favorable for local control (P = 0.046) and was statistically significant for disease-free survival (P = 0.021). The incidence rate of acute and chronic complications between the two groups was also compared. It is possible that the treatment modality enhances local control due to improved conformal dose distributions and the escalated radiation dose applied

  16. Retrospective analysis of role of interstitial brachytherapy using template (MUPIT in locally advanced gynecological malignancies

    Directory of Open Access Journals (Sweden)

    Nandwani Pooja

    2007-01-01

    disease after external beam radiotherapy. Time of gap between EBRT and implant also had an impact on the outcome. Conclusion : Interstitial template brachytherapy by MUPIT is a good alternative to deliver high dose radiation in locally advanced gynecological malignancies where conventional brachytherapy application is either not feasible or likely to give optimal dose distribution. Loco regional control obtained is definitely better than EBRT alone and within the accepted range of complications.

  17. Posttreatment biopsy results following interstitial brachytherapy in early-stage prostate cancer

    International Nuclear Information System (INIS)

    Prestidge, Bradley R.; Hoak, David C.; Grimm, Peter D.; Ragde, Haakon; Cavanagh, William; Blasko, John C.

    1997-01-01

    Purpose: To assess pathologic control rates for prostatic carcinoma as determined by postimplant prostate biopsy in a large series of consecutive patients who have received permanent interstitial brachytherapy using a contemporary transrectal ultrasound-directed, transperineal, computer generated, volume technique. Methods and Materials: Four hundred and two patients received permanent 125 I or 103 Pd interstitial brachytherapy as primary treatment for early stage prostatic carcinoma at the Northwest Tumor Institute between January 1988 and January 1994. Of these, 201 have consented to biopsy 12 or more months postimplant with a median follow-up of 40 months (range: 12-83 months). None had received homonal manipulation. A total of 361 biopsies was performed on 201 patients with a range of one to six annual biopsies per patient (91 received multiple, serial biopsies). Of the 161 patients more than 12 months postimplant who have not been biopsied, most have been unwilling or unable to submit to biopsy. Only six patients with biochemical progression have not been biopsied. There was no difference in the presenting characteristics or implant parameters between those patients biopsied and those that were not. One hundred and forty-three received 125 I (71%) prescribed to a MPD of 160 Gy with a median activity of 35.5 mCi, and 58 (29%) received 103 Pd prescribed to a MPD of 115 Gy with a median activity of 123 mCi. Multiple biopsies were performed under transrectal ultrasound guidance, and all specimens were classified as either negative, indeterminate, or positive. Results: At the time of last biopsy, 161 (80%) have achieved negative pathology, 34 (17%) remain indeterminate, and 6 (3%) have been positive. Only 2 of the 186 patients with a PSA < 4.0 ng/ml at the time of biopsy were positive. Among those 33 indeterminate patients with a subsequent biopsy, 28 have converted to negative, 2 to positive, and 3 remain unchanged to date. Conclusions: These data demonstrate at

  18. Ultrasound-guided interstitial brachytherapy in the treatment of advanced vaginal recurrences from cervical and endometrial carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Weitmann, H.D.; Knocke, T.H.; Waldhaeusl, C.; Poetter, R. [Dept. of Radiotherapy and Radiobiology, Medical Univ. of Vienna (Austria)

    2006-02-01

    Background: in advanced vaginal recurrences of cervical and endometrial carcinomas therapeutic options are rare because of preceding therapy. Patients and methods: 23 patients developing advanced vaginal recurrences of cervical and endometrial carcinomas were included. 15 patients started with external-beam therapy to the pelvis and eight patients after preceding radiotherapy underwent brachytherapy alone. All patients had ultrasound-guided implantation of transvaginal or transperineal interstitial needles for brachytherapy. Median prescribed total dose was 64 Gy. Results: 18 patients (78%) achieved complete remission. Six patients are alive without tumor and one with tumor after a median follow-up of 64 months. 14 patients died of tumor and two of intercurrent disease. 5-year disease-specific survival and local control rate were 43% and 47%, respectively, in patients with complete remission. Univariate analysis found time to relapse > 2 years, initial diameter {<=} 4 cm, initial volume < 15 cm{sup 3}, no extension to the pelvic side wall, volume before brachytherapy < 7.5 cm{sup 3}, brachytherapy coverage index > 0.8, and prescribed total dose > 64 Gy being positive predictors for local control and survival. Conclusion: the use of ultrasound guidance for placement of interstitial needles in template-based brachytherapy of advanced recurrent gynecologic malignancies is a feasible, safe, and cheap method with encouraging results. Today, ultrasound imaging can be also used to some extent for treatment planning which requires further development. Patient- and treatment-related prognostic factors can be defined. (orig.)

  19. Neuro-oncology update: radiation safety and nursing care during interstitial brachytherapy

    International Nuclear Information System (INIS)

    Randall, T.M.; Drake, D.K.; Sewchand, W.

    1987-01-01

    Radiation control and safety are major considerations for nursing personnel during the care of patients receiving brachytherapy. Since the theory and practice of radiation applications are not part of the routine curriculum of nursing programs, the education of nurses and other health care professionals in radiation safety procedures is important. Regulatory agencies recommend that an annual safety course be given to all persons frequenting, using, or associated with patients containing radioactive materials. This article presents pertinent aspects of the principles and procedures of radiation safety, the role of personnel dose-monitoring devices, and the value of additional radiation control features, such as a lead cubicle, during interstitial brain implants. One institution's protocol and procedures for the care of high-intensity iridium-192 brain implants are discussed. Preoperative teaching guidelines and nursing interventions included in the protocol focus on radiation control principles

  20. Technical aspects of the integration of three-dimensional treatment planning dose parameters (GEC-ESTRO Working Group) into pre-implant planning for LDR gynecological interstitial brachytherapy.

    Science.gov (United States)

    Chi, A; Gao, M; Nguyen, N P; Albuquerque, K

    2009-06-01

    This study investigates the technical feasibility of pre-implant image-based treatment planning for LDR GYN interstitial brachytherapy(IB) based on the GEC-ESTRO guidelines. Initially, a virtual plan is generated based on the prescription dose and GEC-ESTRO defined OAR dose constraints with a pre-implant CT. After the actual implant, a regular diagnostic CT was obtained and fused with our pre-implant scan/initial treatment plan in our planning software. The Flexi-needle position changes, and treatment plan modifications were made if needed. Dose values were normalized to equivalent doses in 2 Gy fractions (LQED 2 Gy) derived from the linear-quadratic model with alpha/beta of 3 for late responding tissues and alpha/beta of 10 for early responding tissues. D(90) to the CTV, which was gross tumor (GTV) at the time of brachytherapy with a margin to count for microscopic disease, was 84.7 +/- 4.9% of the prescribed dose. The OAR doses were evaluated by D(2cc) (EBRT+IB). Mean D(2cc) values (LQED(2Gy)) for the rectum, bladder, sigmoid, and small bowel were the following: 63.7 +/- 8.4 Gy, 61.2 +/- 6.9 Gy, 48.0 +/- 3.5 Gy, and 49.9 +/- 4.2 Gy. This study confirms the feasibility of applying the GEC-ESTRO recommended dose parameters in pre-implant CT-based treatment planning in GYN IB. In the process, this pre-implant technique also demonstrates a good approximation of the target volume dose coverage, and doses to the OARs.

  1. Projector-based augmented reality for intuitive intraoperative guidance in image-guided 3D interstitial brachytherapy.

    Science.gov (United States)

    Krempien, Robert; Hoppe, Harald; Kahrs, Lüder; Daeuber, Sascha; Schorr, Oliver; Eggers, Georg; Bischof, Marc; Munter, Marc W; Debus, Juergen; Harms, Wolfgang

    2008-03-01

    The aim of this study is to implement augmented reality in real-time image-guided interstitial brachytherapy to allow an intuitive real-time intraoperative orientation. The developed system consists of a common video projector, two high-resolution charge coupled device cameras, and an off-the-shelf notebook. The projector was used as a scanning device by projecting coded-light patterns to register the patient and superimpose the operating field with planning data and additional information in arbitrary colors. Subsequent movements of the nonfixed patient were detected by means of stereoscopically tracking passive markers attached to the patient. In a first clinical study, we evaluated the whole process chain from image acquisition to data projection and determined overall accuracy with 10 patients undergoing implantation. The described method enabled the surgeon to visualize planning data on top of any preoperatively segmented and triangulated surface (skin) with direct line of sight during the operation. Furthermore, the tracking system allowed dynamic adjustment of the data to the patient's current position and therefore eliminated the need for rigid fixation. Because of soft-part displacement, we obtained an average deviation of 1.1 mm by moving the patient, whereas changing the projector's position resulted in an average deviation of 0.9 mm. Mean deviation of all needles of an implant was 1.4 mm (range, 0.3-2.7 mm). The developed low-cost augmented-reality system proved to be accurate and feasible in interstitial brachytherapy. The system meets clinical demands and enables intuitive real-time intraoperative orientation and monitoring of needle implantation.

  2. Projector-Based Augmented Reality for Intuitive Intraoperative Guidance in Image-Guided 3D Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Krempien, Robert; Hoppe, Harald; Kahrs, Lueder; Daeuber, Sascha; Schorr, Oliver; Eggers, Georg; Bischof, Marc; Munter, Marc W.; Debus, Juergen; Harms, Wolfgang

    2008-01-01

    Purpose: The aim of this study is to implement augmented reality in real-time image-guided interstitial brachytherapy to allow an intuitive real-time intraoperative orientation. Methods and Materials: The developed system consists of a common video projector, two high-resolution charge coupled device cameras, and an off-the-shelf notebook. The projector was used as a scanning device by projecting coded-light patterns to register the patient and superimpose the operating field with planning data and additional information in arbitrary colors. Subsequent movements of the nonfixed patient were detected by means of stereoscopically tracking passive markers attached to the patient. Results: In a first clinical study, we evaluated the whole process chain from image acquisition to data projection and determined overall accuracy with 10 patients undergoing implantation. The described method enabled the surgeon to visualize planning data on top of any preoperatively segmented and triangulated surface (skin) with direct line of sight during the operation. Furthermore, the tracking system allowed dynamic adjustment of the data to the patient's current position and therefore eliminated the need for rigid fixation. Because of soft-part displacement, we obtained an average deviation of 1.1 mm by moving the patient, whereas changing the projector's position resulted in an average deviation of 0.9 mm. Mean deviation of all needles of an implant was 1.4 mm (range, 0.3-2.7 mm). Conclusions: The developed low-cost augmented-reality system proved to be accurate and feasible in interstitial brachytherapy. The system meets clinical demands and enables intuitive real-time intraoperative orientation and monitoring of needle implantation

  3. Interstitial hyperthermia using 8 MHz radiofrequency and stereotaxic brachytherapy for brain tumors

    International Nuclear Information System (INIS)

    Nishimura, Satoshi

    1990-01-01

    As a preliminary study of the interstitial hyperthermia combined with interstitial irradiation (brachytherapy) for the treatment of malignant brain tumors, we performed an experiment of interstitial hyperthermia of brain tissue of dogs. Nine afterloading tubes, four for needle electrodes and five for thermisters, were inserted in the brain tissue of dogs. Rise and stability of temperature were ascertained, and clinical safety was confirmed. Thereafter this combined therapy was applied on seven cases, in which three were malignant gliomas and four were metastatic tumors. Through the guide tubes, 192 Ir thin wires were implanted stereotaxically, and interstitial irradiation was carried out. After removal of 192 Ir wires, needle electrodes were inserted through the same tubes, and also a thermister was guided at the center of electrodes. And interstitial hyperthermia using 8 MHz radiofrequency was carried out. The results of the treatment were evaluated with CT scan based on criteria of the Japan Neurological Society. In cases of malignant gliomas, 2 PRs (partial remission), and 1 NC (no change) were obtained. In cases of metastatic tumors, 1 CR (complete remission), 2 PRs, 1 NC were obtaind. In cases of NCs, progression of tumors have been suppressed for 10 and 17 months, and still alive. As complication, transient worsening of neurological symptoms were observed in four cases (increased paresis: two cases, nausea and vomiting: two cases). The author have had an impression that interstitial hyperthermia combined with interstitial irradiation might become an effective means of treatment of brain tumors. (author)

  4. Outcome of treatment of upper third vaginal recurrences of cervical and endometrial carcinomas with interstitial brachytherapy

    International Nuclear Information System (INIS)

    Charra, C.; Roy, P.; Coquard, R.; Romestaing, P.; Ardiet, J.M.; Gerard, J.P.

    1998-01-01

    Purpose: To describe an original brachytherapy technique using a dedicated intravaginal template for the treatment of vaginal vault recurrences and to evaluate the results of such a treatment. Methods and Materials: Between 1978 and 1993, 78 patients with isolated recurrence of cervical or endometrial carcinoma located in the vaginal vault have been treated in Lyon. Initial treatment was surgery alone in 49 cases and irradiation with surgery in 37 cases. Treatment of the vaginal recurrence was performed with interstitial Iridium 192 brachytherapy combined with pelvic external beam radiation therapy in 34 patients. The tumor was implanted with a dedicated intravaginal plastic template. Six parallel metallic needles were implanted in the vaginal vault and afterloaded with Iridium 192 wires of 4 to 6 cm long. The mucosa of the upper half of the vagina received the same dose as the one encompassing the tumor on the 85% isodose of the Paris system. Results: At 5 years the local control rate was 70% and the overall survival rate 56%. Grade 3 complications occurred in 10% of the cases and only in patients who had received irradiation during the initial treatment of the primary tumor. Conclusions: This brachytherapy technique makes it possible to perform Iridium 192 implants in a difficult situation with a favorable long-term control rate and an acceptable rate of complications

  5. Results in patients treated with high-dose-rate interstitial brachytherapy for oral tongue cancer

    International Nuclear Information System (INIS)

    Yamamoto, Michinori; Shirane, Makoto; Ueda, Tsutomu; Miyahara, Nobuyuki

    2006-01-01

    Eight patients were treated with high-dose-rate interstitial brachytherapy for oral tongue cancer between September 2000 and August 2004. The patient distribution was 1 T1, 5 T2, 1 T3, and 1 T4a. Patients received 50-60 Gy in 10 fractions over seven days with high-dose-rate brachytherapy. Six of the eight patients were treated with a combination of external beam radiotherapy (20-30 Gy) and interstitial brachytherapy. The two-year primary local control rate was 83% for initial case. High-dose-rate brachytherapy was performed safely even for an aged person, and was a useful treatment modality for oral tongue cancer. (author)

  6. A comparison between tandem and ovoids and interstitial gynecologic template brachytherapy dosimetry using a hypothetical computer model

    International Nuclear Information System (INIS)

    Hsu, I-Chow J.; Speight, Joycelyn; Hai, Jenny; Vigneault, Eric; Phillips, Theodore; Pouliot, Jean

    2002-01-01

    Purpose: To evaluate the dose distribution within the clinical target volume between two gynecologic brachytherapy systems - the tandem and ovoids and the Syed-Neblett gynecologic template - using a hypothetical computer model. Methods and Materials: Source positions of an intracavitary system (tandem and ovoids) and an interstitial system (GYN template) were digitized into the Nucletron Brachytherapy Planning System. The GYN template is composed of a 13-catheter implant (12 catheters plus a tandem) based on the Syed-Neblett gynecologic template. For the tandem and ovoids, the dwell times of all sources were evenly weighted to produce a pear-shaped isodose distribution. For the GYN template, the dwell times were determined using volume optimization. The prescribed dose was then normalized to point A in the intracavitary system and to a selected isodose line in the interstitial system. The treated volume in the two systems was kept approximately the same, and a cumulative dose-volume histogram of the treated volume was then generated with the Nucletron Brachytherapy Planning System to use for comparison. To evaluate the dose to a hypothetical target, in this case the cervix, a 2-cm-long, 3-cm-diameter cylinder centered along the tandem was digitized as the clinical target volume. The location of this hypothetical cervix was based on the optimal application of the brachytherapy system. A visual comparison of clinical target coverage by the treated volume on three different orthogonal planes through the treated volume was performed. The percentage dose-volume histograms of the target were generated for comparison. Multiple midline points were also placed at 5-mm intervals away from the tandem in the plane of the cervix to simulate the location of potential bladder and rectal dose points. Doses to these normal structures were calculated for comparison. Results: Although both systems covered the hypothetical cervix adequately, the interstitial system had a better

  7. Time, dose and volume factors in interstitial brachytherapy combined with external irradiation for oral tongue carcinoma

    International Nuclear Information System (INIS)

    Yorozu, Atsunori

    1996-01-01

    This is a retrospective analysis of 136 patients with squamous cell carcinoma of stages I and II of the oral tongue who were treated with interstitial brachytherapy alone or in combination with external irradiation between 1976 and 1991. Control of the primary lesion and the occurrence of late complications were analyzed with respect to dose, time and tumor size with the Cox hazard model. The 5-year survival rates for stages I and II were 84.5% and 75.6%. The 5-year primary control rate was 91.3% for stage I and 77.3% for stage II (p 50 Gy compared with a brachytherapy dose 30 mm. Late complications should be reduced by using a spacer, improvements in dental and oral hygiene, and a sophisticated implant method. (author)

  8. Quality assurance in breast cancer brachytherapy: geographic miss in the interstitial boost treatment of the tumor bed.

    Science.gov (United States)

    Sedlmayer, F; Rahim, H B; Kogelnik, H D; Menzel, C; Merz, F; Deutschmann, H; Kranzinger, M

    1996-03-15

    To assess the role of geographic misses in the interstitial boost treatment of breast cancer patients and to evaluate methods of optimizing breast implants in design, performance, and dosimetry. During lumpectomy, the tumor excision sites of 89 patients were marked by five hemoclips. Postoperative radiographs demonstrated the clips' positions with respect to the extension of the surgical cavity, which was demarcated by air and hematoseroma. Twenty-seven selected patients received interstitial boosts to the tumor bed. The implant was first designed according to the clinical assumptions of the tumor bed's topography and then compared with the radiological findings. Prior to brachytherapy, the planning of the implant's dimension and the needle guidance was performed under simulator control. Dose distributions were first calculated following the Paris System and then electively optimized for the target volume by changing source positions and dwell times. Compared to clinical estimations, the radiological determination of the tumor bed's location revealed an overall potential of topographic errors of 51.8% (14 out of 27 patients), rising up to 78.5% in patients with large adipose breasts (11 out of 13 patients). This observation was due to a high mobility of the tissue, leading to varying tumor site projections at the time of mammography, surgery, and brachytherapy. In all patients, the presimulation of the implant resulted in an adequate coverage of the target volume. In 17 of the 27 treated patients, dose distributions were modified to achieve a higher dose delivery in zones where a higher residual tumor load was expected (boost-in-boost). Breast implants have a high potential of geographic misses that can be avoided by intraoperative clip demarcation. The delineation of the tumor bed allows for dose reports actually referring to the target volume and not to the implant system to be obtained. In addition, modern afterloading techniques offer possibilities of

  9. Quality assurance in breast cancer brachytherapy: geographic miss in the interstitial boost treatment of the tumor bed

    International Nuclear Information System (INIS)

    Sedlmayer, Felix; Rahim, Hassan B. K.; Kogelnik, H. Dieter; Menzel, Christian; Merz, Florian; Deutschmann, Heinz; Kranzinger, Manfred

    1996-01-01

    Purpose: To assess the role of geographic misses in the interstitial boost treatment of breast cancer patients and to evaluate methods of optimizing breast implants in design, performance, and dosimetry. Methods and Materials: During lumpectomy, the tumor excision sites of 89 patients were marked by five hemoclips. Postoperative radiographs demonstrated the clips' positions with respect to the extension of the surgical cavity, which was demarcated by air and hematoseroma. Twenty-seven selected patients received interstitial boosts to the tumor bed. The implant was first designed according to the clinical assumptions of the tumor bed's topography and then compared with the radiological findings. Prior to brachytherapy, the planning of the implant's dimension and the needle guidance was performed under simulator control. Dose distributions were first calculated following the Paris System and then electively optimized for the target volume by changing source positions and dwell times. Results: Compared to clinical estimations, the radiological determination of the tumor bed's location revealed an overall potential of topographic errors of 51.8% (14 out of 27 patients), rising up to 78.5% in patients with large adipose breasts (11 out of 13 patients). This observation was due to a high mobility of the tissue, leading to varying tumor site projections at the time of mammography, surgery, and brachytherapy. In all patients, the presimulation of the implant resulted in an adequate coverage of the target volume. In 17 of the 27 treated patients, dose distributions were modified to achieve a higher dose delivery in zones where a higher residual tumor load was expected (boost-in-boost). Conclusion: Breast implants have a high potential of geographic misses that can be avoided by intraoperative clip demarcation. The delineation of the tumor bed allows for dose reports actually referring to the target volume and not to the implant system to be obtained. In addition, modern

  10. Effect of geometrical optimization on the treatment volumes and the dose homogeneity of biplane interstitial brachytherapy implants

    International Nuclear Information System (INIS)

    Anacak, Yavuz; Esassolak, Mustafa; Aydin, Ayhan; Aras, Arif; Olacak, Ibrahim; Haydaroglu, Ayfer

    1997-01-01

    Background and purpose: The isodose distributions of HDR stepping source brachytherapy implants can be modified by changing dwell times and this procedure is called optimization. The purpose of this study is to evaluate the effect of geometrical optimization on the brachytherapy volumes and the dose homogeneity inside the implant and to compare them with non-optimized counterparts. Material and methods: A set of biplane breast implants consisting of 84 different configurations have been digitized by the planning computer and volumetric analysis was performed for both non-optimized and geometrically optimized implants. Treated length (T L ), treated volume (V 100 ), irradiated volume (V 50 ), overdose volume (V 200 ) and quality index (QI) have been calculated for every non-optimized implant and compared to its corresponding geometrically optimized implant having a similar configuration and covering the same target length. Results: The mean T L was 74.48% of the active length (A L ) for non-optimized implants and was 91.87% for optimized implants (P 50 /V 100 value was 2.71 for non-optimized implants and 2.65 for optimized implants (P 200 /V 100 value was 0.09 for non-optimized implants and 0.10 for optimized implants (P < 0.001). Conclusions: By performing geometrical optimization it is possible to implant shorter needles for a given tumour to adequately cover the target volume with the reference isodose and thus surgical damage is reduced. The amount of healthy tissues outside the target receiving considerable radiation is significantly reduced due to the decrease in irradiated volume. Dose homogeneity inside the implant is significantly improved. Although there is a slight increase of overdose volume inside the implant, this increase is considered to be negligible in clinical applications

  11. Interstitial brachytherapy with 192-IR in treatment of recurrent malignant primary brain tumors. Braquiterapia intersticial con iridio-192 en el tratamiento de recidivas de tumores cerebrales tras cirugia y radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Cardenes, R.; Martinez, R.; Victoria, C.; Nuez, L.; Clavo, B.; Sancedo, G. (Clinica Puerta de Hierro. Madrid (Spain))

    1994-01-01

    Seven patients with recurrent malignant primary brain tumors after surgery and radiation therapy were treated at the Clinica Puerta de Hierro (Madrid) by interstitial brachytherapy with 192-Ir sources. Implantations were performed using computerized tomography and dose prescription were determined following the Paris system rules for interstitial implants. The means dose deliberated was 50 to 65 Gy to the reference isodoses. At the last follow-up all patients except for one are alive and without evidence of progression of the disease. (Author) 35 refs.

  12. Combined transperineal radiofrequency (RF) interstitial hyperthermia and brachytherapy for localized prostate cancer (PC)

    International Nuclear Information System (INIS)

    Urakami, Shinji; Gonda, Nobuko; Kikuno, Nobuyuki

    2001-01-01

    Hyperthermia has been used effectively as a radiation sensitizer. Interstitial hyperthermoradiotherapy has been therefore utilized as a minimal invasive therapy in attempts to improve local tumor control for various cancers, but not for urological cancer. The purpose of this study was to investigate the safety and feasibility of transperineal hyperthermoradiotherapy for localized PC. Based on our basic study of hyperthermoradiotherapy, we devised the procedure of combined transperineal RF interstitial hyperthermia and brachytherapy for localized prostate cancer. Two patients with localized PC underwent transperineal RF interstitial hyperthermia combined with brachytherapy operation the 192-Ir remote after-loading system (RALS). Under transrectal ultrasound guidance, a total number of 12-18 stainless steel needles for 192-Ir RALS were implanted into the prostatic gland and seminal vesicles (SV) in an optimized pattern. Eight of the needles were used as electrodes for hyperthermia, and were electrically insultated using the vinyl catheter along the length of the subdermal fatty tissue to protect from overheating. Three other needles were utilized for continuous temperature mapping in the prostate. Rectal temperature was also monitored. Total radiation doses of 70 Gy to the prostate and SV were planned as a combination of brachytherapy (24 Gy/4 fraction) and external irradiation using a four-field box technique (46 Gy/23 fraction). Hyperthermic treatment (goal of 42 to 43 deg C for 60 minutes) was performed twice following the 1st and 4th brachytherapy at an interval of more than 48 hours for the recovery of cancer cells from thermotolerance. Both patients reached the treatment goal of all intraprostatic temperatures >43.0 deg C, which was considered favorable for hyperthermia, and the rectal temperatures of both patients remained <38 deg C during hyperthermia. In serial PSA measurements of both patients, serum PSA was less than 1.0 ng/ml within 3 months and has since

  13. American brachytherapy society recommends no change for prostate permanent implant dose prescriptions using iodine-125 or palladium-103

    International Nuclear Information System (INIS)

    Rivard, M.J.; Butler, W.M.; Merrick, G.S.; Devlin, P.M.; Hayes, J.K.; Hearn, R.A.; Lief, E.P.; Meigooni, A.S.; Williamson, J.F.

    2008-01-01

    Purpose - In 2004, the American Association of Physicists in Medicine (AAPM) issued a report outlining recommended 125 I and 103 Pd datasets for consistency in calculating brachytherapy dose distributions. In 2005, to aid evaluating the clinical impact of implementing these datasets, the AAPM assessed the historical dependence of how prescribed doses differed from administered doses for 125 I and 103 Pd for permanent implantation of the prostate. Consequently, the American Brachytherapy Society (ABS) considered the nature of these changes towards issuing recommended dose prescriptions for 125 I and 103 Pd interstitial brachytherapy implants for mono-therapy and standard boosts. Methods and materials - An investigation was performed of the 2005 AAPM analysis to determine changes in administered dose while affixing prescribed dose using 2004 AAPM 125 I and 103 Pd brachytherapy dosimetry datasets for prostate implants. For 125 I and 103 Pd, administered dose would change by +1.4% and +4.2%, respectively. The biological and societal impact of changing prescribed dose was considered. Results - Based on the need for clinical constancy and in recognition of overall uncertainties, the ABS recommends immediate implementation of the 2004 AAPM consensus brachytherapy dosimetry datasets and no changes to 125 I and 103 Pd dose prescriptions at this time. Conclusions - Radiation oncologists should continue to prescribe mono-therapy doses of 145 Gy and 125 Gy for 125 I and 105 Pd, respectively, and standard boost doses of 100-110 Gy and 90-100 Gy for 125 I and 103 Pd, respectively. (authors)

  14. Impact of oedema on implant geometry and dosimetry for temporary high dose rate brachytherapy of the prostate

    International Nuclear Information System (INIS)

    Kiffer, J.D.; Schumer, W.A.; Mantle, C.A.; McKenzie, B.J.; Feigen, M.; Quong, G.G.; Waterman, F.M.

    2003-01-01

    The optimal timing of dosimetry for permanent seed prostatic implants remains contentious given the half life of post-implant oedema resolution. The aim of this study was to establish whether prostatic oedematous change over the duration of a temporary high dose rate (HDR) interstitial brachytherapy (BR) boost would result in significant needle displacement, and whether this change in geometry would influence dosimetry. Two CT scans, one for dosimetric purposes on the day of the implant and the second just prior to implant removal, were obtained for four patients receiving transperineal interstitial prostate brachytherapy. The relative changes in cross-sectional dimensions of the implants were calculated by establishing the change in mean radial distance (MRD) of the needle positions from the geometric centre of the implant for each patient's pair of CT studies. The treatment plan, as calculated from the first CT scan, was used in the second set of CT images to allow a comparison of dose distribution. The percentage change in MRD over the duration of the temporary implants ranged from -1.91% to 1.95%. The maximum change in estimated volume was 3.94%. Dosimetric changes were negligible. In the four cases studied, the degree of oedematous change and consequent displacement of flexiguide needle positions was negligible and did not impact on the dosimetry. The rate and direction of oedematous change can be extremely variable but on the basis of the four cases studied and the results of a larger recent study, it might not be necessary to re-image patients for dosimetric purposes over the duration of a fractionated HDR BT boost to the prostate where flexiguide needles are utilized. Nevertheless, further investigation with larger patient numbers is required. Copyright (2003) Blackwell Science Pty Ltd

  15. The American Brachytherapy Society recommendations for low-dose-rate brachytherapy for carcinoma of the cervix

    International Nuclear Information System (INIS)

    Nag, Subir; Chao, Clifford; Erickson, Beth; Fowler, Jeffery; Gupta, Nilendu; Martinez, Alvaro; Thomadsen, Bruce

    2002-01-01

    Purpose: This report presents guidelines for using low-dose-rate (LDR) brachytherapy in the management of patients with cervical cancer. Methods: Members of the American Brachytherapy Society (ABS) with expertise in LDR brachytherapy for cervical cancer performed a literature review, supplemented by their clinical experience, to formulate guidelines for LDR brachytherapy of cervical cancer. Results: The ABS strongly recommends that radiation treatment for cervical carcinoma (with or without chemotherapy) should include brachytherapy as a component. Precise applicator placement is essential for improved local control and reduced morbidity. The outcome of brachytherapy depends, in part, on the skill of the brachytherapist. Doses given by external beam radiotherapy and brachytherapy depend upon the initial volume of disease, the ability to displace the bladder and rectum, the degree of tumor regression during pelvic irradiation, and institutional practice. The ABS recognizes that intracavitary brachytherapy is the standard technique for brachytherapy for cervical carcinoma. Interstitial brachytherapy should be considered for patients with disease that cannot be optimally encompassed by intracavitary brachytherapy. The ABS recommends completion of treatment within 8 weeks, when possible. Prolonging total treatment duration can adversely affect local control and survival. Recommendations are made for definitive and postoperative therapy after hysterectomy. Although recognizing that many efficacious LDR dose schedules exist, the ABS presents suggested dose and fractionation schemes for combining external beam radiotherapy with LDR brachytherapy for each stage of disease. The dose prescription point (point A) is defined for intracavitary insertions. Dose rates of 0.50 to 0.65 Gy/h are suggested for intracavitary brachytherapy. Dose rates of 0.50 to 0.70 Gy/h to the periphery of the implant are suggested for interstitial implant. Use of differential source activity or

  16. Radiological protection on interstitial brachytherapy and dose determination and exposure rate of an Ir-192 source through the MCNP-4B

    International Nuclear Information System (INIS)

    Morales L, M.E.

    2006-01-01

    The present work was carried out in the Neurological Sciences Institute having as objective to determine the dose and the rate of exhibition of the sources of Iridium 192, Iodine 125 and Palladium 103; which are used to carry out implant in the Interstitial Brachytherapy according to the TG43. For it we carry out a theoretical calculation, its are defined in the enter file: the geometry, materials of the problem and the radiation source, etc; in the MCNP-4B Monte Carlo code, considering a punctual source and for the dose determination we simulate thermoluminescent dosemeters (TLD): at 5 cm, 50 cm, 100 cm and 200 cm of the source. Our purpose is to analyze the radioprotection measures that should take into account in this Institute in which are carried out brain biopsies using a Micro mar stereotactic mark, and in a near future with the collaboration of a doctor and a cuban physique seeks to be carried out the Interstitial Brachytherapy technique with sources of Ir-192 for patient with tumors like glioblastoma, astrocytoma, etc. (Author)

  17. Natural History of Clinically Staged Low- and Intermediate-Risk Prostate Cancer Treated With Monotherapeutic Permanent Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Taira, Al V.; Merrick, Gregory S.; Galbreath, Robert W.; Wallner, Kent E.; Butler, Wayne M.

    2010-01-01

    Purpose: To evaluate the natural history of clinically staged low- and intermediate-risk prostate cancer treated with permanent interstitial seed implants as monotherapy. Methods and Materials: Between April 1995 and May 2005, 463 patients with clinically localized prostate cancer underwent brachytherapy as the sole definitive treatment. Men who received supplemental external beam radiotherapy or androgen deprivation therapy were excluded. Dosimetric implant quality was determined based on the minimum dose that covered 90% of the target volume and the volume of the prostate gland receiving 100% of the prescribed dose. Multiple parameters were evaluated as predictors of treatment outcomes. Results: The 12-year biochemical progression-free survival (bPFS), cause-specific survival, and overall survival rates for the entire cohort were 97.1%, 99.7%, and 75.4%, respectively. Only pretreatment prostate-specific antigen level, percent positive biopsy cores, and minimum dose that covered 90% of the target volume were significant predictors of biochemical recurrence. The bPFS, cause-specific survival, and overall survival rates were 97.4%, 99.6%, and 76.2%, respectively, for low-risk patients and 96.4%, 100%, and 74.0%, respectively, for intermediate-risk patients. The bPFS rate was 98.8% for low-risk patients with high-quality implants versus 92.1% for those with less adequate implants (p < 0.01), and it was 98.3% for intermediate-risk patients with high-quality implants versus 86.4% for those with less adequate implants (p < 0.01). Conclusions: High-quality brachytherapy implants as monotherapy can provide excellent outcomes for men with clinically staged low- and intermediate-risk prostate cancer. For these men, a high-quality implant can achieve results comparable to high-quality surgery in the most favorable pathologically staged patient subgroups.

  18. The relative efficacy of HDR and LDR interstitial brachytherapy in squamous cell carcinoma of vagina

    International Nuclear Information System (INIS)

    Demanes, D. J.; Hsu, I-C.; Lin, S.; Ewing, T.; Rodriguez, R.

    1996-01-01

    Introduction: Beginning in 1982 we performed low dose rate (LDR) interstitial template brachytherapy (ISTB) for carcinoma of the vagina. High dose rate (HDR) remote afterloading has been used exclusively since 1991. We compare the results LDR and HDR brachytherapy. Material and Methods: Between 1982 and 1994, 30 patients with primary squamous cell carcinoma of vagina received external beam radiotherapy (EBRT) and brachytherapy. The AJCC stage distribution was 3 stage I, 25 stage II, and 2 stage III. The average central pelvic EBRT dose was 35 Gy. Pelvic side wall EBRT doses ranged from 45 to 50.4 Gy. Nineteen patients had LDR treatment; 3 intracavitary brachytherapy (ICB) and 16 ISTB. Eleven patients had HDR treatment; 2 ICB and 9 ISB. The average dose delivered by LDR was 41.2 Gy usually in 2 fractions, and by HDR 32.5 Gy in 6 fractions of 500-550 cGy. Local failures were confirmed pathologically. The absolute survival (AS) and relapse-free survival (RFS) were calculated using Kaplan-Meier method and compared with logrank statistics. Results: The mean follow-up was 77 months for LDR and 23 months for HDR. Local and regional control was achieved in 90% (27/30) of the patients. Three year AS was 84% and RFS was 87%. There was no significant difference between LDR and HDR in AS, RFS or local-regional control, (log rank p=0.85, p=0.12 and p=0.35 respectively). The single HDR local failure presented in a patient with extensive stage II disease who declined ISTB. There were fewer complications following HDR. The 1 case of extensive vaginal necrosis and the 3 cases of rectovaginal fistula that required surgery occurred only with LDR brachytherapy. Discussion: Excellent local and regional control of carcinoma of the vagina can be achieved by administering limited doses of external radiation and brachytherapy. Interstitial template implants are the best means of encompassing paravaginal disease while sparing the adjacent uninvolved normal tissues from high doses of

  19. Interstitial high-dose-rate brachytherapy in the treatment of base of tongue carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Takacsi-Nagy, Z.; Polgar, C.; Somogyi, A.; Major, T.; Fodor, J.; Nemeth, G. [Dept. of Radiotherapy, National Inst. of Oncology, Budapest (Hungary); Oberna, F. [Dept. of Maxillofacial Surgery, St. Rokus Hospital, Budapest (Hungary); Remenar, E.; Kasler, M. [Dept. of Head and Neck, Maxillofacial and Reconstructive Plastic Surgery, National Inst. of Oncology, Budapest (Hungary)

    2004-12-01

    Background and purpose: to date none of the studies examined the feasibility and efficacy of interstitial high-dose-rate (HDR) brachytherapy in the treatment of carcinoma of the tongue base. Therefore the aim of this study was to contribute to this issue. Patients and methods: between 1992 and 2000 37 patients (mean age 55 years) with T1-4 and NO-3 carcinoma of the base of tongue were presented. Neck dissection was carried out in twelve cases (32%). 30 patients with advanced stage received brachytherapy boost after 50-66.5 Gy (mean, 60 Gy) locoregional external beam irradiation (EBI) and 7 patients with early stage (T1-2, NO) were managed locally with wide tumor excision and sole brachytherapy. 4 of them underwent neck dissection and the others were subjected to 50 Gy regional EBI. The mean dose of boost and sole brachytherapy was 18 Gy and 28 Gy, respectively. Results: the median follow-up time for surviving patients was 51 months. The 7 sole brachytherapy patients are living with no evidence of disease. For patients treated with EBI and brachytherapy boost, the 5-year actuarial rate of local, locoregional recurrence-free and overall survival was 60%, 52% and 46%, respectively. For all patients in univariate analysis larger tumor size (T4 vs. T1-3) was significant negative predictor of local (RR: 7.23) and locoregional control (RR: 3.87), but nodal involvement was not. Delayed soft tissue ulceration and osteoradionecrosis occurred in 4 (13%) EBI and brachytherapy treated patients. None of the sole brachytherapy patients experienced severe late radiation toxicity. Conclusion: EBI combined with interstitial HDR brachytherapy boost result in acceptable local tumor control with low incidence of late side effects in patients with advanced disease. Fractionated sole HDR brachytherapy following tumor excision is a feasible treatment option for patients with early stage cancer and gives excellent local results. (orig.)

  20. Electromagnetic tracking for treatment verification in interstitial brachytherapy

    DEFF Research Database (Denmark)

    Bert, Christoph; Kellermeier, Markus; Tanderup, Kari

    2016-01-01

    Electromagnetic tracking (EMT) is used in several medical fields to determine the position and orientation of dedicated sensors, e.g., attached to surgical tools. Recently, EMT has been introduced to brachytherapy for implant reconstruction and error detection. The manuscript briefly summarizes...

  1. Physics and quality assurance for brachytherapy - Part II: Low dose rate and pulsed dose rate

    International Nuclear Information System (INIS)

    Williamson, Jeffrey F.

    1997-01-01

    Purpose: A number of recent developments have revitalized brachytherapy including remote afterloading, implant optimization, increasing use of 3D imaging, and advances in dose specification and basic dosimetry. However, the core physical principles underlying the classical methods of dose calculation and arrangement of multiple sources remain unchanged. The purpose of this course is to review these principles and their applications to low dose-rate interstitial and intracavitary brachytherapy. Emphasis will be placed upon the classical implant systems along with classical and modern methods of dose specification. The level of presentation is designed for radiation oncology residents and beginning clinical physicists. A. Basic Principles (1) Radium-substitute vs. low-energy sealed sources (2) Dose calculation principles (3) The mysteries of source strength specification revealed: mgRaEq, mCi and air-kerma strength B. Interstitial Brachytherapy (1) Target volume, implanted volume, dose specification in implants and implant optimization criteria (2) Classical implant systems: Manchester Quimby and Paris a) Application of the Manchester system to modern brachytherapy b) Comparison of classical systems (3) Permanent interstitial implants a) Photon energy and half life b) Dose specification and pre-operative planning (4) The alphabet soup of dose specification: MCD (mean central dose), minimum dose, MPD (matched peripheral dose), MPD' (minimum peripheral dose) and DVH (dose-volume histogram) quality indices C. Intracavitary Brachytherapy for Carcinoma of the Cervix (1) Basic principles a) Manchester System: historical foundation of U.S. practice patterns b) Principles of applicator design (2) Dose specification and treatment prescription a) mg-hrs, reference points, ICRU Report 38 reference volume -- Point A dose vs mg-hrs and IRAK (Integrated Reference Air Kerma) -- Tissue volume treated vs mg-hrs and IRAK b) Practical methods of treatment specification and prescription

  2. Physics and quality assurance for brachytherapy - Part II: Low dose rate and pulsed dose rate

    International Nuclear Information System (INIS)

    Williamson, Jeffrey F.

    1996-01-01

    Purpose: A number of recent developments have revitalized brachytherapy including remote afterloading, implant optimization, increasing use of 3D imaging, and advances in dose specification and basic dosimetry. However, the core physical principles underlying the classical methods of dose calculation and arrangement of multiple sources remain unchanged. The purpose of this course is to review these principles and their applications to low dose-rate interstitial and intracavitary brachytherapy. Emphasis will be placed upon the classical implant systems along with classical and modern methods of dose specification. The level of presentation is designed for radiation oncology residents and beginning clinical physicists. A. Basic Principles (1) Radium-substitute vs. low-energy sealed sources (2) Dose calculation principles (3) The mysteries of source strength specification revealed: mgRaEq, mCi and air-kerma strength B. Interstitial Brachytherapy (1) Target volume, implanted volume, dose specification in implants and implant optimization criteria (2) Classical implant systems: Manchester Quimby and Paris a) Application of the Manchester system to modern brachytherapy b) Comparison of classical systems (3) Permanent interstitial implants a) Photon energy and half life b) Dose specification and pre-operative planning (4) The alphabet soup of dose specification: MCD (mean central dose), minimum dose, MPD (matched peripheral dose), MPD' (minimum peripheral dose) and DVH (dose-volume histogram) quality indices C. Intracavitary Brachytherapy for Carcinoma of the Cervix (1) Basic principles a) Manchester System: historical foundation of U.S. practice patterns b) Principles of applicator design (2) Dose specification and treatment prescription a) mg-hrs, reference points, ICRU Report 38 reference volume --Point A dose vs mg-hrs and IRAK (Integrated Reference Air Kerma) --Tissue volume treated vs mg-hrs and IRAK b) Practical methods of treatment specification and prescription

  3. Dose-volume analysis for quality assurance of interstitial brachytherapy for breast cancer

    International Nuclear Information System (INIS)

    Vicini, Frank A.; Kestin, Larry L.; Edmundson, Gregory K.; Jaffray, David A.; Wong, John W.; Kini, Vijay R.; Chen, Peter Y.; Martinez, Alvaro A.

    1999-01-01

    Purpose/Objective: The use of brachytherapy in the management of breast cancer has increased significantly over the past several years. Unfortunately, few techniques have been developed to compare dosimetric quality and target volume coverage concurrently. We present a new method of implant evaluation that incorporates computed tomography-based three-dimensional (3D) dose-volume analysis with traditional measures of brachytherapy quality. Analyses performed in this fashion will be needed to ultimately assist in determining the efficacy of breast implants. Methods and Materials: Since March of 1993, brachytherapy has been used as the sole radiation modality after lumpectomy in selected protocol patients with early-stage breast cancer treated with breast-conserving therapy. Eight patients treated with high-dose-rate (HDR) brachytherapy who had surgical clips outlining the lumpectomy cavity and underwent computed tomography (CT) scanning after implant placement were selected for this study. For each patient, the postimplant CT dataset was transferred to a 3D treatment planning system. The lumpectomy cavity, target volume (lumpectomy cavity plus a 1-cm margin), and entire breast were outlined on each axial slice. Once all volumes were entered, the programmed HDR brachytherapy source positions and dwell times were imported into the 3D planning system. Using the tools provided by the 3D planning system, the implant dataset was then registered to the visible implant template in the CT dataset. The distribution of the implant dose was analyzed with respect to defined volumes via dose-volume histograms (DVH). Isodose surfaces, the dose homogeneity index, and dosimetric coverage of the defined volumes were calculated and contrasted. All patients received 32 Gy to the entire implanted volume in 8 fractions of 4 Gy over 4 days. Results: Three-plane implants were used for 7 patients and a two-plane implant for 1 patient. The median number of needles per implant was 16.5 (range

  4. Clinical Use of the Utrecht Applicator for Combined Intracavitary/Interstitial Brachytherapy Treatment in Locally Advanced Cervical Cancer

    International Nuclear Information System (INIS)

    Nomden, Christel N.; Leeuw, Astrid A.C. de; Moerland, Marinus A.; Roesink, Judith M.; Tersteeg, Robbert J.H.A.; Jürgenliemk-Schulz, Ina Maria

    2012-01-01

    Purpose: The aims of this study were to investigate the benefit of the Utrecht interstitial CT/MR applicator for combined intracavitary/interstitial (IC/IS) approach, using magnetic resonance imaging—guided brachytherapy, over the intracavitary approach alone in patients with locally advanced cervical cancer and to analyze the clinical use of needles. Methods and Materials: This study includes the first 20 patients treated with the new applicator. Brachytherapy consisted of two pulsed dose rate applications, and the second application was performed with the IC/IS approach. The number of needles, chosen guiding holes through the ovoids, and insertion depths were based on the dose distribution and dosimetric shortcomings of the first application (IC alone). We investigated the dosimetric gain by comparing the clinical interstitial optimized plan (IC/IS clinical ) with an additionally generated optimized plan without needle use (IC study ). Furthermore, we studied the relation of the inserted needles and their source loading patterns with the high-risk clinical target volume (HR-CTV). Results: A total of 54 needles (range, 1–6 per application) were applied with an average depth of 25 mm. The chosen needle positions corresponded with the location of the HR-CTV extensions. The total and individual needle treatment times per application were on average 19% (range, 4–35%) and 7% (range, 2–14%) of the implant treatment time, respectively. The total (external-beam radiotherapy + brachytherapy) D90 HR-CTV for the IC study and the IC/IS clinical were on average 79.5 (SD 7.4) Gy α/β10 and 83.9 (SD 6.7) Gy α/β10 , respectively, with an average gain of 4.4 (SD 2.3) Gy α/β10 for the second application. Conclusions: Needle placement was feasible in all patients and resulted in a gain in dose and better coverage of HR-CTV. Defining the location of HR-CTV protrusions and analyzing the associated needles has given us deeper understanding of the possibilities in magnetic

  5. SU-F-T-55: Reproducibility of Interstitial HDR Brachytherapy Plans

    Energy Technology Data Exchange (ETDEWEB)

    Lee, S; Ellis, R; Traughber, B; Podder, T [University Hospitals Case Medical Center, Cleveland, OH (United States)

    2016-06-15

    Purpose: Treating gynecological cancers with interstitial high-dose-rate (HDR) brachytherapy requires precise reconstruction of catheter positions to obtain accurate dosimetric plans. In this study, we investigated the degree of reproducibility of dosimetric plans for Syed HDR brachytherapy. Methods: We randomly selected five patients having cervix-vaginal cancer who were recently treated in our clinic with interstitial HDR brachytherapy with a prescription dose of 25–30 Gy in five fractions. Interstitial needles/catheters were placed under fluoroscopic guidance and intra-operative 3T MRI scan was performed to confirm the desired catheter placement for adequate target volume coverage. A CT scan was performed and fused with the MRI for delineating high-risk CTV (HR-CTV), intermediate-risk CTV (IR-CTV) and OARs. HDR treatment plans were generated using Oncentra planning software. A single plan was used for all five fractions of treatment for each patient. For this study, we took the original clinical plan and removed all the reconstructed catheters from the plan keeping the original contours unchanged. Then, we manually reconstructed all the catheters and entered the same dwell time from the first original clinical plan. The dosimetric parameters studied were: D90 for HR-CTV and IR-CV, and D2cc for bladder, rectum, sigmoid and bowel. Results: The mean of absolute differences in dosimetric coverage (D90) were (range): 1.3% (1.0–2.0%) and 2.0% (0.9–3.6%) for HR-CTV and IR-CTV, respectively. In case of OARs, the mean of absolute variations in D2cc were (range): 4.7% (0.7–8.9%) for bladder, 1.60% (0.3–3.2%) for rectum, 1.6% (0–3.9%) for sigmoid, and 1.8% (0–5.1%) for bowel. Conclusion: Overall, the reproducibility of interstitial HDR plans was within clinically acceptable limit. Observed maximum variation in D2cc for bladder. If number of catchers and dwell points were relatively low or any one catheter was heavily loaded, then reproducibility of the plan

  6. The use of disposable vascular catheters in interstitial brachytherapy of skin cancers

    International Nuclear Information System (INIS)

    Daly, N.J.; Malissard, L.; Douchez, J.; Combes, P.F.

    1978-01-01

    Technical improvements dealing with interstitial brachytherapy (Ir 192 ) of skin cancers are presented. Fine disposable plastic tubes fitted with mandril, which allow loading of light radioactive material in any case are used. Short term results are discussed according to 101 applications [fr

  7. Interstitial brachytherapy for eyelid carcinoma. Outcome analysis in 60 patients

    Energy Technology Data Exchange (ETDEWEB)

    Krengli, M.; Deantonio, L. [University Hospital ' ' Maggiore della Carita' ' , Division of Radiotherapy, Novara (Italy); University of ' ' Piemonte Orientale' ' , Department of Translational Medicine, Novara (Italy); Masini, L.; Filomeno, A.; Gambaro, G. [University Hospital ' ' Maggiore della Carita' ' , Division of Radiotherapy, Novara (Italy); Comoli, A.M. [University Hospital Maggiore della Carita, Ophthalmology, Novara (Italy); Negri, E. [University Hospital Maggiore della Carita, Medical Physics, Novara (Italy)

    2014-03-15

    Eyelid cancer is a therapeutic challenge due to the cosmetic and functional implications of this anatomical region and the objectives of therapy are tumor control, functional and cosmetic outcome. The present study was performed to analyze local control, toxicity, functional and cosmetic results in patients with eyelid carcinoma treated by interstitial brachytherapy. In this study 60 patients with eyelid carcinoma were treated by interstitial brachytherapy using iridium ({sup 192}Ir) wires with a linear activity of 1.2-1.7 mCi/cm. The prescription dose was 51-70 Gy (mean 65 Gy, median 66 Gy). Of the 60 patients 51 (85.0 %) had received no prior treatment, 4 (6.7 %) had received previous surgery with positive or close margins and 5 (8.3 %) had suffered local recurrence after surgery. Of the tumors 52 (86.7 %) were basal cell carcinoma, 7 (11.7 %) squamous cell carcinoma and 1 (1.7 %) Merkel cell carcinoma. Clinical stage of the 51 previously untreated tumors was 38 T1N0, 12 T2N0 and 1 T3N0. Mean follow-up was 92 months (range 6-253 months). Local control was maintained in 96.7 % of patients. Late effects higher than grade 2 were observed in 3.0 % of cases. Functional and cosmetic outcomes were optimal in 68.4 % of patients. Interstitial brachytherapy for carcinoma of the eyelid can achieve local control, cosmetic and functional results comparable to those of surgery. (orig.) [German] Das Karzinom des Augenlids stellt aufgrund der funktionellen und kosmetischen Beeintraechtigungen dieser anatomischen Region eine therapeutische Herausforderung dar. Ziele der Therapie sind sowohl die Tumorkontrolle als auch ein gutes funktionelles und kosmetisches Ergebnis. Lokale Kontrolle, Toxizitaet sowie funktionelle und kosmetische Ergebnisse bei Patienten mit Karzinom des Augenlids, die mit interstitieller Brachytherapie behandelt wurden, sollten analysiert werden. Sechzig Patienten mit Karzinom des Augenlids wurden mit interstitieller Brachytherapie mit Iridium-192-Draehten

  8. Dose rate constant and energy spectrum of interstitial brachytherapy sources

    International Nuclear Information System (INIS)

    Chen Zhe; Nath, Ravinder

    2001-01-01

    In the past two years, several new manufacturers have begun to market low-energy interstitial brachytherapy seeds containing 125 I and 103 Pd. Parallel to this development, the National Institute of Standards and Technology (NIST) has implemented a modification to the air-kerma strength (S K ) standard for 125 I seeds and has also established an S K standard for 103 Pd seeds. These events have generated a considerable number of investigations on the determination of the dose rate constants (Λ) of interstitial brachytherapy seeds. The aim of this work is to study the general properties underlying the determination of Λ and to develop a simple method for a quick and accurate estimation of Λ. As the dose rate constant of clinical seeds is defined at a fixed reference point, we postulated that Λ may be calculated by treating the seed as an effective point source when the seed's source strength is specified in S K and its source characteristics are specified by the photon energy spectrum measured in air at the reference point. Using a semi-analytic approach, an analytic expression for Λ was derived for point sources with known photon energy spectra. This approach enabled a systematic study of Λ as a function of energy. Using the measured energy spectra, the calculated Λ for 125 I model 6711 and 6702 seeds and for 192 Ir seed agreed with the AAPM recommended values within ±1%. For the 103 Pd model 200 seed, the agreement was 5% with a recently measured value (within the ±7% experimental uncertainty) and was within 1% with the Monte Carlo simulations. The analytic expression for Λ proposed here can be evaluated using a programmable calculator or a simple spreadsheet and it provides an efficient method for checking the measured dose rate constant for any interstitial brachytherapy seed once the energy spectrum of the seed is known

  9. Assessment of dose homogeneity in conformal interstitial breast brachytherapy with special respect to ICRU recommendations

    Directory of Open Access Journals (Sweden)

    Tibor Major

    2011-09-01

    Full Text Available Purpose: To present the results of dose homogeneity analysis for breast cancer patients treated with image-basedconformal interstitial brachytherapy, and to investigate the usefulness of the ICRU recommendations. Material and methods: Treatment plans of forty-nine patients who underwent partial breast irradiation with interstitialbrachytherapy were analyzed. Quantitative parameters were used to characterize dose homogeneity. Dose nonuniformityratio (DNR, dose homogeneity index (DHI, uniformity index (UI and quality index (QI were calculated.Furthermore, parameters recommended by the ICRU 58 such as minimum target dose (MTD, mean central dose (MCD,high dose volume, low dose volume and the spread between local minimum doses were determined. Correlationsbetween the calculated homogeneity parameters and usefulness of the ICRU parameters in image-based brachytherapywere investigated. Results: Catheters with mean number of 15 (range: 6-25 were implanted in median 4 (range: 3-6 planes. The volu -me of the PTV ranged from 15.5 cm3 to 176 cm3. The mean DNR was 0.32, the DHI 0.66, the UI 1.49 and the QI 1.94. Relatedto the prescribed dose, the MTD was 69% and the MCD 135%. The mean high dose volume was 8.1 cm3 (10%, whilethe low dose volume was 63.8 cm3 (96%. The spread between minimum doses in central plane ranged from –14% to+20%. Good correlation was found between the DNR and the DHI (R2 = 0.7874, and the DNR correlated well with theUI (R2 = 0.7615 also. No correlation was found between the ICRU parameters and any other volumetric parameters. Conclusions: To characterize the dose uniformity in high-dose rate breast implants, DVH-related homogeneityparameters representing the full 3D dose distributions are mandatory to be used. In many respects the current re commendationsof the ICRU Report 58 are already outdated, and it is well-timed to set up new recommendations, whichare more feasible for image-guided conformal interstitial brachytherapy.

  10. Treatment planning for multicatheter interstitial brachytherapy of breast cancer – from Paris system to anatomy-based inverse planning

    Directory of Open Access Journals (Sweden)

    Tibor Major

    2017-02-01

    Full Text Available In the last decades, treatment planning for multicatheter interstitial breast brachytherapy has evolved considerably from fluoroscopy-based 2D to anatomy-based 3D planning. To plan the right positions of the catheters, ultrasound or computed tomography (CT imaging can be used, but the treatment plan is always based on postimplant CT images. With CT imaging, the 3D target volume can be defined more precisely and delineation of the organs at risk volumes is also possible. Consequently, parameters calculated from dose-volume histogram can be used for quantitative plan evaluation. The catheter reconstruction is also easier and faster on CT images compared to X-ray films. In high dose rate brachytherapy, using a stepping source, a number of forward dose optimization methods (manual, geometrical, on dose points, graphical are available to shape the dose distribution to the target volume, and these influence dose homogeneities to different extent. Currently, inverse optimization algorithms offer new possibilities to improve dose distributions further considering the requirements for dose coverage, dose homogeneity, and dose to organs at risk simultaneously and automatically. In this article, the evolvement of treatment planning for interstitial breast implants is reviewed, different forward optimization methods are discussed, and dose-volume parameters used for quantitative plan evaluation are described. Finally, some questions of the inverse optimization method are investigated and initial experiences of the authors are presented.

  11. Reirradiation for recurrent head and neck cancer with salvage interstitial pulsed-dose-rate brachytherapy. Long-term results

    Energy Technology Data Exchange (ETDEWEB)

    Strnad, Vratislav; Lotter, Michael; Kreppner, Stephan; Fietkau, Rainer [University Hospital Erlangen, Dept. of Radiation Oncology, Erlangen (Germany)

    2015-01-10

    To assess the long-term results of protocol-based interstitial pulsed-dose-rate (PDR) brachytherapy as reirradiation combined with simultaneous chemotherapy and interstitial hyperthermia in selected patients with recurrent head and neck tumors. A total of 104 patients with biopsy-proven recurrent head and neck cancer were treated with interstitial PDR brachytherapy. Salvage surgery had also been undergone by 53/104 (51 %) patients (R1 or R2 resection in > 80 % of patients). Salvage brachytherapy alone was administered in 81 patients (78 %), with a median total dose of 56.7 Gy. Salvage brachytherapy in combination with external beam radiotherapy (EBRT) was performed in 23/104 patients (32 %), using a median total dose of D{sub REF} = 24 Gy. Simultaneously to PDR brachytherapy, concomitant chemotherapy was administered in 58/104 (55.8 %) patients. A single session of interstitial hyperthermia was also used to treat 33/104 (31.7 %) patients. The analysis was performed after a median follow-up of 60 months. Calculated according to Kaplan-Meier, local tumor control rates after 2, 5, and 10 years were 92.5, 82.4, and 58.9 %, respectively. Comparing results of salvage PDR brachytherapy with or without simultaneous chemotherapy, the 10-year local control rates were 76 vs. 39 % (p= 0014), respectively. No other patient- or treatment-related parameters had a significant influence on treatment results. Soft tissue necrosis or bone necrosis developed in 18/104 (17.3 %) and 11/104 (9.6 %) patients, respectively, but only 3 % of patients required surgical treatment. PDR interstitial brachytherapy with simultaneous chemotherapy is a very effective and, in experienced hands, also a safe treatment modality in selected patients with head and neck cancer in previously irradiated areas. (orig.) [German] Es erfolgte die Analyse der Langzeitergebnisse einer protokollbasierten interstitiellen Brachytherapie (Re-Bestrahlung) mit simultaner Chemotherapie und interstitieller Hyperthermie

  12. Modeling of interstitial diffusion of ion-implanted boron

    International Nuclear Information System (INIS)

    Velichko, O.I.; Knyazheva, N.V.

    2009-01-01

    A model of the interstitial diffusion of ion-implanted boron during rapid thermal annealing of silicon layers previously amorphized by implantation of germanium has been proposed. It is supposed that the boron interstitials are created continuously during annealing due to generation, dissolution, or rearrangement of the clusters of impurity atoms which are formed in the ion-implanted layers with impurity concentration above the solubility limit. The local elastic stresses arising due to the difference of boron atomic radius and atomic radius of silicon also contribute to the generation of boron interstitials. A simulation of boron redistribution during thermal annealing for 60 s at a temperature of 850 C has been carried out. The calculated profile agrees well with the experimental data. A number of the parameters of interstitial diffusion have been derived. In particular, the average migration length of nonequilibrium boron interstitials is equal to 12 nm. It was also obtained that approximately 1.94% of boron atoms were converted to the interstitial sites, participated in the fast interstitial migration, and then became immobile again transferring into a substitutional position or forming the electrically inactive complexes with crystal lattice defects. (authors)

  13. Introduction of a hybrid treatment delivery system used for quality assurance in multi-catheter interstitial brachytherapy

    Science.gov (United States)

    Kallis, Karoline; Kreppner, Stephan; Lotter, Michael; Fietkau, Rainer; Strnad, Vratislav; Bert, Christoph

    2018-05-01

    Multi-catheter interstitial brachytherapy (iBT) is a treatment option for breast cancer patients after breast conserving surgery. Typically, only a few additional quality interventions after the first irradiation have been introduced to ensure the planned treatment delivery. Therefore, the purpose of this study is to show the possibilities of an electromagnetic tracking (EMT) system integrated into the afterloader for quality assurance (QA) in high-dose rate (HDR) iBT of patients with breast cancer. The hybrid afterloader system equipped with an electromagnetic sensor was used for all phantom and patient measurements. Phantom measurements were conducted to estimate the quality of different evaluation schemes. After a coherent point drift registration of the EMT traces to the reconstructed catheters based on computed tomograms the dwell positions (DP) were defined. Different fitting and interpolation methods were analyzed for the reconstruction of DPs. All estimated DPs were compared to the DPs defined in treatment planning. Until now, the implant geometry of 20 patients treated with HDR brachytherapy was acquired and explored. Regarding the reconstruction techniques, both fitting and interpolation were able to detect manually introduced shifts and swaps. Nonetheless, interpolation showed superior results (RMSE  =  1.27 mm), whereas fitting seemed to be more stable to distortion and motion. The EMT system proved to be beneficial for QA in brachytherapy and furthermore, clinical feasibility was proven.

  14. Interstitial radiotherapy

    International Nuclear Information System (INIS)

    Scardino, P.T.; Bretas, F.

    1987-01-01

    The authors now have 20 years of experience with modern techniques of brachytherapy. The large number of patients treated in medical centers around the world and the widespread use of this type of radiotherapy have provided us with substantial information about the indications and contraindications, advantages and disadvantages, pitfalls and complications, as well as the results of these techniques. Although the focus of this review is the experience at Baylor using the combined technique of gold seed implantation plus external beam irradiation, the alternative forms of brachytherapy will be described and compared. The authors' intention is to provide the busy clinician with a succinct and informative review indicating the status of modern interstitial radiotherapy and describing day-to-day approach and results

  15. Computed tomographic-guided iodine-125 interstitial implants for malignant thoracic tumors

    Energy Technology Data Exchange (ETDEWEB)

    Huang, Qiming [The Department of Radiology, First Affiliated Hospital of Fujian Medical University, 20 Chazhong Road, Fuzhou 350005 (China); The Department of Radiology, Second Affiliated Hospital of Fujian Medical University, 34 Zhongshan Bei Road, Quanzhou 362000 (China); Chen, Jin; Chen, Qunlin [The Department of Radiology, First Affiliated Hospital of Fujian Medical University, 20 Chazhong Road, Fuzhou 350005 (China); Lai, Qingquan; Cai, Siqing [The Department of Radiology, Second Affiliated Hospital of Fujian Medical University, 34 Zhongshan Bei Road, Quanzhou 362000 (China); Luo, Kaidong [The Department of Radiology, Longyan Hosptial of Traditional Chinese Medical, 59 Longteng Middle Road, Longyan 364000 (China); Lin, Zhengyu, E-mail: linsinlan@yahoo.com.cn [The Department of Radiology, First Affiliated Hospital of Fujian Medical University, 20 Chazhong Road, Fuzhou 350005 (China)

    2013-11-01

    Purpose: To evaluate the feasibility and efficacy of percutaneous interstitial brachytherapy using iodine-125 ({sup 125}I) radioactive seeds under computed tomographic (CT) guidance for malignant thoracic tumors. Materials and methods: Forty-one patients (34 males, 7 females; 18–90 years; mean, 63.7 years) with 77 lesions (3 in the mediastinum, 7 in the chest wall, 67 in the lung) underwent percutaneous interstitial implantation of {sup 125}I radioactive seeds under CT guidance. A treatment planning system (TPS) was employed to calculate the number and distribution of seeds preoperatively. An 18-G needle was inserted into the lesions under CT guidance and send the seeds according to TPS. Two patients with mediastinal lesions undergoing seed implantation received an artificial pneumothorax. One patient with lung carcinoma adjacent to the anterior mediastinum underwent seed implantation through the sternum. Follow-up CT was done every 2 months postoperatively. Results: The procedure was successful in all patients. No major procedure-associated death occurred. The mean duration of follow-up was 19.4 ± 1.3 months (3–49 months). A complete response (CR) was seen in 49 lesions (63.6%), partial response (PR) in 9 lesions (11.7%), stable disease (SD) in 12 lesions (12.8%), and progressive disease (PD) in 7 lesions (7.4%). The overall response rate (CR + PR) was 75.3%; the local control rate (CR + PR + SD) was 90.9%. The 1-, 2- and 3-year progression-free rates for local tumors were 91%, 88% and 88%, respectively. The 1-, 2- and 3-year survival rates were 87%, 74% and 68%, respectively. Conclusion: Implantation of CT-guided {sup 125}I seeds is feasible and effective for patients with malignant thoracic tumors.

  16. Indications and technical aspects of brachytherapy in breast conserving treatment of breast cancer

    International Nuclear Information System (INIS)

    Erik Van, Limbergen

    2003-01-01

    Improved local control rates have been demonstrated in retrospective studies as well as in randomized trials on brachytherapy with increasing doses to the tumour bed. The higher local control obtained by interstitial breast implants, as compared to external photon or electron beam boosts, have been mainly attributed to the higher doses actually delivered to the tumour bed by these implants for the same nominal dose as compared to external beam radiotherapy (RT). On the other hand, poor cosmesis has also been correlated with radiation dose to the breast skin (radiation telangiectases), and breast tissue (retraction due to fibrosis), the latter depending not only on RT dose but also on the treated boost volume. For this reason, a possible benefit of interstitial implants will only be realized when the gain in local control goes together with minimal cosmetic damage. Therefore, the ballistic advantages of interstitial implants have to be maximally exploited: i.e. the treated volume should be maximally adapted to the target volume, and additional irradiation of the breast skin by the boost technique should be avoided. This paper deals in detail with the technical aspects of breast brachytherapy that seem to be relevant for high quality outcome. (author)

  17. Interstitial irradiation of rectal carcinoma with rectal template

    International Nuclear Information System (INIS)

    Ikeda, Hiroshi; Nishiyama, Kinji; Tanaka, Ken; Nakanishi, Makoto; Inoue, Takehiro

    1984-01-01

    Using Iridium-192 wires through a rectal template after Syed, interstitial brachytherapy was conducted in a patient with inoperable adenocarcinoma of the rectum. 67-year-old man with constipation and change in the stool caliber underwent external radiotherapy (4,000cGy/4W) to the whole pelvis including the perineum, followed by interstitial implant using a template, at the Department of Radiology, Osaka University Hospital. Marked tumor regression, marked circumferential fibrosis and a remarkable decline of CEA titers (pre-RT: 35.8ng/ml, post-RT: 6.2ng/ml) were observed until 7 months post-RT. The domestic production of Iridium-192 wires has made possible the intergrated use of brachytherapy in the perineal region in Japan. (author)

  18. Influence of 125I seed interstitial brachytherapy on recovery of facial nerve function

    International Nuclear Information System (INIS)

    Song Tieli; Zheng Lei; Zhang Jie; Cai Zhigang; Yang Zhaohui; Yu Guangyan; Zhang Jianguo

    2010-01-01

    Objective: To study the influence of 125 I seed interstitial brachytherapy in parotid region on the recovery of facial nerve function. Methods: A total of the data of 21 patients with primary parotid carcinoma were treated with resection and 125 I interstitial brachytherapy. All the patients had no facial palsy before operation and the prescribed dose was 60 Gy. During 4 years of follow-up, the House-Brackmann grading scales and ENoG were used to evaluate the function of facial nerve. According to the modified regional House-Brackmann grading scales, the facial nerve branches of patients in affected side were divided into normal and abnormal groups, and were compared with those in contra-lateral side. Results: Post-operation facial palsy occurred in all the patients, but the facial palsy recovered within 6 months. The latency time differences between affected side and contralateral side were statistically significant in abnormal group from 1 week to 6 months after treatment (t=2.362, P=0.028), and were also different in normal group 1 week after treatment (t=2.522, P=0.027). Conclusions: 125 I interstitital brachytherapy has no influence on recovery of facial nerve function after tumor resection and no delayed facial nerve damage. (authors)

  19. Salvage high-dose-rate interstitial brachytherapy for locally recurrent rectal cancer

    Energy Technology Data Exchange (ETDEWEB)

    Pellizzon, Antonio Cassio Assis, E-mail: acapellizzon@hcancer.org.br [A.C. Camargo Cancer Center, Sao Paulo, SP (Brazil). Departamento de Radioterapia

    2016-05-15

    For tumors of the lower third of the rectum, the only safe surgical procedure is abdominal-perineal resection. High-dose-rate interstitial brachytherapy is a promising treatment for local recurrence of previously irradiated lower rectal cancer, due to the extremely high concentrated dose delivered to the tumor and the sparing of normal tissue, when compared with a course of external beam radiation therapy. (author)

  20. Post-treatment biopsy results following permanent transrectal ultrasound-guided interstitial brachytherapy in early stage prostate cancer

    International Nuclear Information System (INIS)

    Prestidge, Bradley R.; Blasko, John C.; Grimm, Peter D.; Hoak, David C.; Cavanagh, Bill; Ragde, Haakon

    1995-01-01

    PURPOSE/OBJECTIVE: Although some controversy remains, most authors agree that post-treatment prostatic biopsy is the best measure of local control in prostate cancer. Brachytherapy series reporting post-implant biopsy results have been few in number, limited in size, and involving older open or combined external beam techniques. The present study was undertaken to assess local control rates as determined by post-implant prostate biopsy in a large series of consecutive patients who have received permanent interstitial brachytherapy using a contemporary transrectal ultrasound directed, transperineal, computer generated, volume technique. METHOD/MATERIALS: From January 1988 to January 1994, 402 patients received permanent I-125 (285, 71%) or Pd-103 (117, 29%) interstitial brachytherapy as primary treatment for prostatic carcinoma at the Northwest Tumor Institute. Of these, 201 have consented to prostatic biopsy at least 12 months post-implant with a median follow-up of 40 months (range of 12 to 83 months). None had received hormone manipulation. A total of 361 biopsies were performed on 201 patients with a range of 1 to 6 yearly biopsies per patient; 91 receiving multiple biopsies. The other 201 patients were either unable (for geographic reasons) or unwilling to submit for biopsy. However, all patients with a rising PSA or clinical suspicion of recurrence underwent biopsy when possible. The 201 biopsy patients presented with a median age of 69 (range 47 to 89). Stages included 51 T1 (25%), 125 T2a (62%), 22 T2b (11%), and 3 T2c (1%). Gleason sums included 69 2-4 (34%), 117 5-6 (58%), 15 7-10 (7%), and 2 ungraded (1%). The initial PSA was 6.6 (range 0.7 to 74.6). There was no significant difference in the presenting characteristics or implant parameters between those patients biopsied and those that were not. 143 received I-125 (71%) prescribed to a minimum peripheral dose of 160 Gy with a median activity of 35.5 mCi, and 58 (29%) received Pd-103 prescribed to a

  1. Predictors of Toxicity After Image-guided High-dose-rate Interstitial Brachytherapy for Gynecologic Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Larissa J. [Department of Radiation Oncology, Brigham and Women' s Hospital and Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts (United States); Viswanathan, Akila N., E-mail: aviswanathan@lroc.harvard.edu [Department of Radiation Oncology, Brigham and Women' s Hospital and Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts (United States)

    2012-12-01

    Purpose: To identify predictors of grade 3-4 complications and grade 2-4 rectal toxicity after three-dimensional image-guided high-dose-rate (HDR) interstitial brachytherapy for gynecologic cancer. Methods and Materials: Records were reviewed for 51 women (22 with primary disease and 29 with recurrence) treated with HDR interstitial brachytherapy. A single interstitial insertion was performed with image guidance by computed tomography (n = 43) or magnetic resonance imaging (n = 8). The median delivered dose in equivalent 2-Gy fractions was 72.0 Gy (45 Gy for external-beam radiation therapy and 24 Gy for brachytherapy). Toxicity was reported according to the Common Toxicity Criteria for Adverse Events. Actuarial toxicity estimates were calculated by the Kaplan-Meier method. Results: At diagnosis, the median patient age was 62 years and the median tumor size was 3.8 cm. The median D90 and V100 were 71.4 Gy and 89.5%; the median D2cc for the bladder, rectum, and sigmoid were 64.6 Gy, 61.0 Gy, and 52.7 Gy, respectively. The actuarial rates of all grade 3-4 complications at 2 years were 20% gastrointestinal, 9% vaginal, 6% skin, 3% musculoskeletal, and 2% lymphatic. There were no grade 3-4 genitourinary complications and no grade 5 toxicities. Grade 2-4 rectal toxicity was observed in 10 patients, and grade 3-4 complications in 4; all cases were proctitis with the exception of 1 rectal fistula. D2cc for rectum was higher for patients with grade 2-4 (68 Gy vs 57 Gy for grade 0-1, P=.03) and grade 3-4 (73 Gy vs 58 Gy for grade 0-2, P=.02) rectal toxicity. The estimated dose that resulted in a 10% risk of grade 2-4 rectal toxicity was 61.8 Gy (95% confidence interval, 51.5-72.2 Gy). Discussion: Image-guided HDR interstitial brachytherapy results in acceptable toxicity for women with primary or recurrent gynecologic cancer. D2cc for the rectum is a reliable predictor of late rectal complications. Three-dimensional-based treatment planning should be performed to ensure

  2. Dose reduction in LDR brachytherapy by implanted prostate gold fiducial markers.

    Science.gov (United States)

    Landry, Guillaume; Reniers, Brigitte; Lutgens, Ludy; Murrer, Lars; Afsharpour, Hossein; de Haas-Kock, Danielle; Visser, Peter; van Gils, Francis; Verhaegen, Frank

    2012-03-01

    The dosimetric impact of gold fiducial markers (FM) implanted prior to external beam radiotherapy of prostate cancer on low dose rate (LDR) brachytherapy seed implants performed in the context of combined therapy was investigated. A virtual water phantom was designed containing a single FM. Single and multi source scenarios were investigated by performing Monte Carlo dose calculations, along with the influence of varying orientation and distance of the FM with respect to the sources. Three prostate cancer patients treated with LDR brachytherapy for a recurrence following external beam radiotherapy with implanted FM were studied as surrogate cases to combined therapy. FM and brachytherapy seeds were identified on post implant CT scans and Monte Carlo dose calculations were performed with and without FM. The dosimetric impact of the FM was evaluated by quantifying the amplitude of dose shadows and the volume of cold spots. D(90) was reported based on the post implant CT prostate contour. Large shadows are observed in the single source-FM scenarios. As expected from geometric considerations, the shadows are dependent on source-FM distance and orientation. Large dose reductions are observed at the distal side of FM, while at the proximal side a dose enhancement is observed. In multisource scenarios, the importance of shadows appears mitigated, although FM at the periphery of the seed distribution caused underdosage (LDR brachytherapy seed implant dose distributions. Therefore, reduced tumor control could be expected from FM implanted in tumors, although our results are too limited to draw conclusions regarding clinical significance.

  3. Method of localization and implantation of the lumpectomy site for high dose rate brachytherapy after conservative surgery for T1 and T2 breast cancer

    International Nuclear Information System (INIS)

    Perera, F.; Chisela, F.; Engel, J.; Venkatesan, V.

    1995-01-01

    Purpose: This article describes our technique of localization and implantation of the lumpectomy site of patients with T1 and T2 breast cancer. Our method was developed as part of our Phase I/II pilot study of high dose rate (HDR) brachytherapy alone after conservative surgery for early breast cancer. Methods and Materials: In March 1992, we started a pilot study of HDR brachytherapy to the lumpectomy site as the sole radiotherapy after conservative surgery for clinical T1 or T2 invasive breast cancer. Initially, the protocol required intraoperative placement of the interstitial needles at the time of definitive surgery to the breast. The protocol was then generalized to allow the implantation of the lumpectomy site after definitive surgery to the breast, either at the time of subsequent axillary nodal dissection or postoperatively. To date, five patients have been implanted intraoperatively at the time of definitive breast surgery. Twelve patients were implanted after definitive breast surgery, with 7 patients being done at the time of axillary nodal dissection and 5 patients postoperatively. We devised a method of accurately localizing and implanting the lumpectomy site after definitive breast surgery. The method relies on the previous placement of surgical clips by the referring surgeon to mark the lumpectomy site. For each patient, a breast mold is made with radio-opaque angiocatheters taped onto the mold in the supero-inferior direction. A planning CT scan is then obtained through the lumpectomy site. The volume of the lumpectomy site, the number of implant planes necessary, and the orientation of the implants are then determined from the CT scan. The angiocatheters provide a reference grid on the CT films to locate the entry and exit points of the interstitial needles on the plastic mold. The entry and exit points for reference needles are then transferred onto the patient's skin enabling implantation of the lumpectomy site. Needle positions with respect to

  4. Role of interstitial implantation in gynecological cancer

    International Nuclear Information System (INIS)

    Nori, D.; Hilaris, B.S.

    1987-01-01

    Recurrent cancer at any site carries a gloomy prognosis. Cancer of the cervix that recurs after radical surgery or curative radiation therapy is a perplexing problem confronting both gynecological and radiation oncologists. In the authors' series, 45% of the patients survived disease-free at 1 year and 10% survived without disease at 5 years or longer following interstitial implantation for recurrent cervical cancer. The optimal utilization of this procedure seems to depend on the site of recurrence, the extent of the disease in the pelvis, and the status of para-aortic node involvement. This retrospective analysis enabled the authors to identify the prognostic factors. The most favorable group benefited by this technique were those who presented with either central recurrence or unilateral, localized pelvic side wall recurrent disease. The least morbidity was noticed in those patients with minimal surgical manipulations at the time of the interstitial implantation. The authors recommended that only a limited and essential surgical procedure should accompany interstitial implantation, since the associated morbidity and mortality is high and survival brief

  5. Evaluation of an active magnetic resonance tracking system for interstitial brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Wang, Wei, E-mail: wwang21@partners.org [Department of Radiology, Brigham and Women’s Hospital, Boston, Massachusetts 02115 and Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women’s Hospital, Boston, Massachusetts 02115 (United States); Viswanathan, Akila N.; Damato, Antonio L.; Cormack, Robert A. [Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women’s Hospital, Boston, Massachusetts 02115 (United States); Chen, Yue; Tse, Zion [Department of Engineering, The University of Georgia, Athens, Georgia 30602 (United States); Pan, Li [Siemens Healthcare USA, Baltimore, Maryland 21287 (United States); Tokuda, Junichi; Schmidt, Ehud J. [Department of Radiology, Brigham and Women’s Hospital, Boston, Massachusetts 02115 (United States); Seethamraju, Ravi T. [Siemens Healthcare USA, Boston, Massachusetts 02115 (United States); Dumoulin, Charles L. [Radiology, Cincinnati Children’s Hospital Medical Center, Cincinnati, Ohio 45229 (United States)

    2015-12-15

    -resolution 3D MRI of the phantom was performed followed by catheter digitization based on the catheter’s imaging artifacts. The catheter trajectory error was characterized in terms of the mean distance between corresponding dwell points in MRTR-generated catheter trajectory and MRI-based catheter digitization. The MRTR-based catheter trajectory reconstruction process was also performed on three gynecologic cancer patients, and then compared with catheter digitization based on MRI and CT. Results: The catheter tip localization error increased as the MRTR stylet moved further off-center and as the stylet’s orientation deviated from the main magnetic field direction. Fifteen catheters’ trajectories were reconstructed by MRTR. Compared with MRI-based digitization, the mean 3D error of MRTR-generated trajectories was 1.5 ± 0.5 mm with an in-plane error of 0.7 ± 0.2 mm and a tip error of 1.7 ± 0.5 mm. MRTR resolved ambiguity in catheter assignment due to crossed catheter paths, which is a common problem in image-based catheter digitization. In the patient studies, the MRTR-generated catheter trajectory was consistent with digitization based on both MRI and CT. Conclusions: The MRTR system provides accurate catheter tip localization and trajectory reconstruction in the MR environment. Relative to the image-based methods, it improves the speed, safety, and reliability of the catheter trajectory reconstruction in interstitial brachytherapy. MRTR may enable in-procedural dosimetric evaluation of implant target coverage.

  6. Evaluation of an active magnetic resonance tracking system for interstitial brachytherapy

    International Nuclear Information System (INIS)

    Wang, Wei; Viswanathan, Akila N.; Damato, Antonio L.; Cormack, Robert A.; Chen, Yue; Tse, Zion; Pan, Li; Tokuda, Junichi; Schmidt, Ehud J.; Seethamraju, Ravi T.; Dumoulin, Charles L.

    2015-01-01

    -resolution 3D MRI of the phantom was performed followed by catheter digitization based on the catheter’s imaging artifacts. The catheter trajectory error was characterized in terms of the mean distance between corresponding dwell points in MRTR-generated catheter trajectory and MRI-based catheter digitization. The MRTR-based catheter trajectory reconstruction process was also performed on three gynecologic cancer patients, and then compared with catheter digitization based on MRI and CT. Results: The catheter tip localization error increased as the MRTR stylet moved further off-center and as the stylet’s orientation deviated from the main magnetic field direction. Fifteen catheters’ trajectories were reconstructed by MRTR. Compared with MRI-based digitization, the mean 3D error of MRTR-generated trajectories was 1.5 ± 0.5 mm with an in-plane error of 0.7 ± 0.2 mm and a tip error of 1.7 ± 0.5 mm. MRTR resolved ambiguity in catheter assignment due to crossed catheter paths, which is a common problem in image-based catheter digitization. In the patient studies, the MRTR-generated catheter trajectory was consistent with digitization based on both MRI and CT. Conclusions: The MRTR system provides accurate catheter tip localization and trajectory reconstruction in the MR environment. Relative to the image-based methods, it improves the speed, safety, and reliability of the catheter trajectory reconstruction in interstitial brachytherapy. MRTR may enable in-procedural dosimetric evaluation of implant target coverage

  7. Perioperative interstitial brachytherapy for recurrent keloid scars

    International Nuclear Information System (INIS)

    Rio, E.; Bardet, E.; Peuvrel, P.; Martinet, L.; Perrot, P.; Baraer, F.; Loirat, Y.; Sartre, J.Y.; Malard, O.; Ferron, C.; Dreno, B.

    2010-01-01

    Purpose: Evaluation of the results of perioperative interstitial brachytherapy with low dose-rate (L.D.R.) Ir-192 in the treatment of keloid scars. Patients and methods: We performed a retrospective analysis of 73 histologically confirmed keloids (from 58 patients) resistant to medico surgical treated by surgical excision plus early perioperative brachytherapy. All lesions were initially symptomatic. Local control was evaluated by clinical evaluation. Functional and cosmetic results were assessed in terms of patient responses to a self-administered questionnaire. Results: Median age was 28 years (range 13-71 years). Scars were located as follows: 37% on the face, 32% on the trunk or abdomen, 16% on the neck, and 15% on the arms or legs. The mean delay before loading was four hours (range, 1-6 h). The median dose was 20 Gy (range, 15-40 Gy). Sixty-four scars (from 53 patients) were evaluated. Local control was 86% (follow-up, 44.5 months; range, 14-150 months). All relapses occurred early within 2 years posttreatment. At 20 months, survival without recurrence was significantly lower when treated lengths were more than 6 cm long. The rate was 100% for treated scars below 4.5 cm in length, 95% (95% CI: 55-96) for those 4.5-6 cm long, and 75% (95% CI: 56-88) beyond 6 cm (p = 0.038). Of the 35 scars (28 patients) whose results were reassessed, six remained symptomatic and the esthetic results were considered to be good in 51% (18/35) and average in 37% (13/35) (median follow-up, 70 months; range, 16-181 months). Conclusion: Early perioperative L.D.R. brachytherapy delivering 20 Gy at 5 mm reduced the rate of recurrent keloids resistant to other treatments and gave good functional results. (authors)

  8. A comparison study on various low energy sources in interstitial prostate brachytherapy.

    Science.gov (United States)

    Bakhshabadi, Mahdi; Ghorbani, Mahdi; Khosroabadi, Mohsen; Knaup, Courtney; Meigooni, Ali S

    2016-02-01

    Low energy sources are routinely used in prostate brachytherapy. (125)I is one of the most commonly used sources. Low energy (131)Cs source was introduced recently as a brachytherapy source. The aim of this study is to compare dose distributions of (125)I, (103)Pd, and (131)Cs sources in interstitial brachytherapy of prostate. ProstaSeed (125)I brachytherapy source was simulated using MCNPX Monte Carlo code. Additionally, two hypothetical sources of (103)Pd and (131)Cs were simulated with the same geometry as the ProstaSeed (125)I source, while having their specific emitted gamma spectra. These brachytherapy sources were simulated with distribution of forty-eight seeds in a phantom including prostate. The prostate was considered as a sphere with radius of 1.5 cm. Absolute and relative dose rates were obtained in various distances from the source along the transverse and longitudinal axes inside and outside the tumor. Furthermore, isodose curves were plotted around the sources. Analyzing the initial dose profiles for various sources indicated that with the same time duration and air kerma strength, (131)Cs delivers higher dose to tumor. However, relative dose rate inside the tumor is higher and outside the tumor is lower for the (103)Pd source. The higher initial absolute dose in cGy/(h.U) of (131)Cs brachytherapy source is an advantage of this source over the others. The higher relative dose inside the tumor and lower relative dose outside the tumor for the (103)Pd source are advantages of this later brachytherapy source. Based on the total dose the (125)I source has advantage over the others due to its longer half-life.

  9. Iodine-125 thin seeds decrease prostate swelling during transperineal interstitial permanent prostate brachytherapy

    International Nuclear Information System (INIS)

    Beydoun, Nadine; Bucci, Joseph A.; Chin, Yaw S.; Malouf, David

    2014-01-01

    Prostate swelling following seed implantation is a well-recognised phenomenon. The purpose of this intervention was to assess whether using thinner seeds reduces post-implant swelling with permanent prostate brachytherapy. Eighteen consecutive patients eligible for prostate seed brachytherapy underwent seed implantation using iodine-125 (I-125) thin seeds. Operative time, dosimetry, prostate swelling and toxicity were assessed and compared with standard I-125 stranded seed controls, sourced from the department's brachytherapy database. A learning curve was noted with the thin seeds in terms of greater bending and deviation of needles from their intended path. This translated into significantly longer total operative time (88 vs 103 minutes; P=0.009, 95% confidence interval (CI) 4.1-24.3) and time per needle insertion (2.6 vs 3.7 minutes; P<0.001, 95% CI 0.5-1.3) for the thin seeds. Day 30 prostate volumes were significantly smaller in the thin seed group compared with standard seeds (40.9cc vs 46.8cc; P=0.001, 95% CI 1.5-5.6). The ratio of preoperative transrectal ultrasound to day 30 post-implant CT volume was also smaller in the thin seed group (1.2±0.1 for standard seeds vs 1.1±0.1 for thin seeds). Post-implant dosimetric parameters were comparable for both groups. No significant differences were seen in acute urinary morbidity or quality of life between the two groups. I-125 thin seeds are associated with an initial learning curve, with longer operative time, even for experienced brachytherapists. The significant reduction in day 30 prostate volumes with the thin seeds has useful implications in terms of optimising dose coverage to the prostate in the early period post-implantation, as well as improving the accuracy of post-implant dosimetric assessments.

  10. Incidence of seed migration to the chest, abdomen, and pelvis after transperineal interstitial prostate brachytherapy with loose 125I seeds

    International Nuclear Information System (INIS)

    Sugawara, Akitomo; Shigematsu, Naoyuki; Nakashima, Jun; Kunieda, Etsuo; Nagata, Hirohiko; Mizuno, Ryuichi; Seki, Satoshi; Shiraishi, Yutaka; Kouta, Ryuichi; Oya, Mototsugu

    2011-01-01

    The aim was to determine the incidence of seed migration not only to the chest, but also to the abdomen and pelvis after transperineal interstitial prostate brachytherapy with loose 125 I seeds. We reviewed the records of 267 patients who underwent prostate brachytherapy with loose 125 I seeds. After seed implantation, orthogonal chest radiographs, an abdominal radiograph, and a pelvic radiograph were undertaken routinely to document the occurrence and sites of seed migration. The incidence of seed migration to the chest, abdomen, and pelvis was calculated. All patients who had seed migration to the abdomen and pelvis subsequently underwent a computed tomography scan to identify the exact location of the migrated seeds. Postimplant dosimetric analysis was undertaken, and dosimetric results were compared between patients with and without seed migration. A total of 19,236 seeds were implanted in 267 patients. Overall, 91 of 19,236 (0.47%) seeds migrated in 66 of 267 (24.7%) patients. Sixty-nine (0.36%) seeds migrated to the chest in 54 (20.2%) patients. Seven (0.036%) seeds migrated to the abdomen in six (2.2%) patients. Fifteen (0.078%) seeds migrated to the pelvis in 15 (5.6%) patients. Seed migration occurred predominantly within two weeks after seed implantation. None of the 66 patients had symptoms related to the migrated seeds. Postimplant prostate D90 was not significantly different between patients with and without seed migration. We showed the incidence of seed migration to the chest, abdomen and pelvis. Seed migration did not have a significant effect on postimplant prostate D90

  11. Indication of brachytherapy of prostate with permanent implants

    International Nuclear Information System (INIS)

    Chauveinc, L.; Solignac, S.; Rosenwald, J.C.; Firmin, F.; Cosset, J.M.; Flam, T.; Thiounn, N.

    2002-01-01

    In the last decade, brachytherapy emerged as a particularly appealing new way of treating localized prostate cancer. Recently published 10-12 years biochemical control results appear to be superimposable to the best percentages achieved by surgery or conformal radiotherapy, with a small percentage of complications. This applied to severely patients. Only patients with T1/T2, PSA 60 g, hip mobility limitations, a urinary obstructive syndrome and previous trans-urethral resection lead to difficulties in technical implantation and therefore must be taken into account when discussing brachytherapy. In conclusion, for adequately selected patients, brachytherapy offers a particularly applied alternative to surgery and external radiotherapy, with satisfactory long term biochemical control rates and limited complications. (author)

  12. A comparison study on various low energy sources in interstitial prostate brachytherapy

    Directory of Open Access Journals (Sweden)

    Mahdi Bakhshabadi

    2016-02-01

    Full Text Available Purpose: Low energy sources are routinely used in prostate brachytherapy. 125 I is one of the most commonly used sources. Low energy 131 Cs source was introduced recently as a brachytherapy source. The aim of this study is to compare dose distributions of 125 I, 103 Pd, and 131 Cs sources in interstitial brachytherapy of prostate. Material and methods: ProstaSeed 125 I brachytherapy source was simulated using MCNPX Monte Carlo code. Additionally, two hypothetical sources of 103 Pd and 131 Cs were simulated with the same geometry as the ProstaSeed 125 I source, while having their specific emitted gamma spectra. These brachytherapy sources were simulated with distribution of forty-eight seeds in a phantom including prostate. The prostate was considered as a sphere with radius of 1.5 cm. Absolute and relative dose rates were obtained in various distances from the source along the transverse and longitudinal axes inside and outside the tumor. Furthermore, isodose curves were plotted around the sources. Results : Analyzing the initial dose profiles for various sources indicated that with the same time duration and air kerma strength, 131 Cs delivers higher dose to tumor. However, relative dose rate inside the tumor is higher and outside the tumor is lower for the 103 Pd source. Conclusions : The higher initial absolute dose in cGy/(h.U of 131 Cs brachytherapy source is an advantage of this source over the others. The higher relative dose inside the tumor and lower relative dose outside the tumor for the 103 Pd source are advantages of this later brachytherapy source. Based on the total dose the 125 I source has advantage over the others due to its longer half-life.

  13. Monte Carlo-aided dosimetry of the new Bebig IsoSeed registered 103Pd Interstitial Brachytherapy Seed

    International Nuclear Information System (INIS)

    Daskalov, George M.; Williamson, J.F.

    2001-01-01

    A new model 103 Pd interstitial brachytherapy source, the IsoSeed registered 103 Pd, was recently introduced by Bebig Isotopentechnik und Umweltdiagnostik GmbH for permanent implant applications. This study presents the first quantitative theoretical study of the seed's dosimetric quantities. Monte Carlo photon transport (MCPT) simulation techniques have been used to evaluate the dose-rate distributions around the model IsoSeed registered 103 Pd source in liquid water and air phantoms. These results have been used to calculate and tabulate the anisotropy function, F(r,θ), radial dose function, g(r), and anisotropy factors, φ(r), and dose-rate constant as defined by AAPM Task Group 43 (TG-43) Report. Cartesian 'away' and 'along' tables, giving the dose rates per unit air-kerma strength in water in the range 0.1-3 cm distance around the seed have also been tabulated. The dose-rate constant, Λ, was evaluated by simulating the wide-angle, free-air chamber (WAFAC) calibration geometry recently implemented by NIST (National Institute of Standards and Technology) to realize the primary standard of air-kerma strength (S K,N99 ) for low-energy photon-emitting brachytherapy sources. The dose-rate constant has been found to be Λ=0.660±0.017 in units of dose-rate per unit air-kerma strength (cGy·h-1·U-1)

  14. Treatment of 'inoperable' neck nodes using surgical clearance and postoperative interstitial irradiation

    International Nuclear Information System (INIS)

    Stafford, N.; Dearnaley, D.

    1988-01-01

    The authors describe their experience with postoperative interstitial irradiation (brachytherapy) using after-loaded iridium-192 ( 192 Ir) wires positioned at the site of the tumour after its surgical exposure and debulking or clearance. All the treated patients had previously received external beam irradiation to the neck and in the first three cases local skin necrosis occurred as a sequel to interstitial implantation. Therefore, in the other five patients, cutaneous or myocutaneous flaps were used to resurface the area at the time of implant placement. This combination of techniques, not described previously, provided local tumour control in four out of the five cases without irradiation necrosis occurring in any of them. (author)

  15. Individualised 3D printed vaginal template for MRI guided brachytherapy in locally advanced cervical cancer

    DEFF Research Database (Denmark)

    Lindegaard, Jacob Christian; Lænsø Madsen, Mads; Hansen, Anders Traberg

    2016-01-01

    Intracavitary–interstitial applicators for MRI guided brachytherapy are becoming increasingly important in locally advanced cervical cancer. The 3D printing technology enables a versatile method for obtaining a high degree of individualisation of the implant. Our clinical workflow is presented...

  16. Pulsed dose rate (PDR) brachytherapy as salvage treatment of locally advanced or recurrent gynecologic cancer

    DEFF Research Database (Denmark)

    Jensen, P T; Roed, H; Engelholm, S A

    1998-01-01

    PURPOSE: Pulsed dose rate (PDR) brachytherapy is a new treatment option permitting dose distribution optimization in interstitial implants. It possesses the advantage of equipment simplification and radiation protection to the staff, compared to the manually afterloading technique. This study pre...

  17. Importance of brachytherapy technique in the management of primary carcinoma of the vagina

    International Nuclear Information System (INIS)

    Stock, R.G.; Mychalczak, B.; Armstrong, J.G.; Hoskins, W.; Harrison, L.B.

    1991-01-01

    Primary vaginal carcinoma is a rare malignancy. There is little information regarding the optimal treatment. Management has primarily been with external-beam radiation therapy and brachytherapy. This paper examines the importance of brachytherapy and the significance of its techniques in the treatment of this disease. Brachytherapy plays an important part in the management of primary vaginal carcinoma. External-beam radiation therapy alone is not an adequate treatment for this disease. For stages II and III disease, there is a trend toward improved disease-free survival with the use of a temporary interstitial implant compared to an intracavitary application

  18. Dosimetric perturbations of a lead shield for surface and interstitial high-dose-rate brachytherapy

    International Nuclear Information System (INIS)

    Candela-Juan, Cristian; Granero, Domingo; Vijande, Javier; Ballester, Facundo; Perez-Calatayud, Jose; Rivard, Mark J

    2014-01-01

    In surface and interstitial high-dose-rate brachytherapy with either 60 Co, 192 Ir, or 169 Yb sources, some radiosensitive organs near the surface may be exposed to high absorbed doses. This may be reduced by covering the implants with a lead shield on the body surface, which results in dosimetric perturbations. Monte Carlo simulations in Geant4 were performed for the three radionuclides placed at a single dwell position. Four different shield thicknesses (0, 3, 6, and 10 mm) and three different source depths (0, 5, and 10 mm) in water were considered, with the lead shield placed at the phantom surface. Backscatter dose enhancement and transmission data were obtained for the lead shields. Results were corrected to account for a realistic clinical case with multiple dwell positions. The range of the high backscatter dose enhancement in water is 3 mm for 60 Co and 1 mm for both 192 Ir and 169 Yb. Transmission data for 60 Co and 192 Ir are smaller than those reported by Papagiannis et al (2008 Med. Phys. 35 4898–4906) for brachytherapy facility shielding; for 169 Yb, the difference is negligible. In conclusion, the backscatter overdose produced by the lead shield can be avoided by just adding a few millimetres of bolus. Transmission data provided in this work as a function of lead thickness can be used to estimate healthy organ equivalent dose saving. Use of a lead shield is justified. (paper)

  19. Study on interstitial brachytherapy using 103Pd seeds on tumor-bearing rats

    International Nuclear Information System (INIS)

    Feng Huiru; Zhang Jingming; Tian Jiahe; Ding Weimin; Bai Hongsheng; Jin Xiaohai

    2003-01-01

    The effects of low-dose-rate brachytherapy are investigated in tumor-bearing rat. Walker 256 cells are transplanted subcutaneously with a trocar in the left leg of rats (Wistar). Two weeks later, rats with a tumor of 10 mm in mean diameter are divided into three groups (10 per group). Two groups are given 1 seed and 2 seeds implantation of 103 Pd, respectively, the third group is as an untreated control. Tumor size is measured twice a week until the 25th day when the rats are killed. Tumor is monitored either by palpation or further confirmed by histopathology. Kaplan-Meier statistic method is performed for survival analysis. The results show that the average weight of rats in untreated group is lower than in radiation groups (P 0.05). Tumor volumes in all treatment groups increase more obviously than in control till 16 days post-implantation. Tumor regression rate in 1 seed group is higher than in control group and in 2 seeds group. Although survival analysis show that the median survival time in 1 seed, 2 seeds and control groups are 24±0, 21±2 and 19±2 days with survival rate of 80%, 60% and 50% respectively, no significant differences are seen in all groups. So, brachytherapy with 103 Pd seed is effective on tumor-bearing rats. The implantation of seed can cause tumor edema in a self-limited way. A reasonable doses chosen for brachytherapy may play a role in treatment success

  20. Accelerated Partial Breast Irradiation With Low-Dose-Rate Interstitial Implant Brachytherapy After Wide Local Excision: 12-Year Outcomes From a Prospective Trial

    Energy Technology Data Exchange (ETDEWEB)

    Hattangadi, Jona A. [Harvard Radiation Oncology Program, Boston, MA (United States); Powell, Simon N. [Department of Radiation Oncology, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); MacDonald, Shannon M.; Mauceri, Thomas; Ancukiewicz, Marek [Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA (United States); Freer, Phoebe [Department of Radiology, Massachusetts General Hospital, Boston, MA (United States); Lawenda, Brian [21st Century Oncology, Las Vegas, NV (United States); Alm El-Din, Mohamed A. [Department of Radiation Oncology, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); Department of Clinical Oncology, Tanta University Hospital, Tanta (Egypt); Gadd, Michele A.; Smith, Barbara L. [Department of Surgical Oncology, Massachusetts General Hospital, Boston, MA (United States); Taghian, Alphonse G., E-mail: ataghian@partners.org [Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA (United States)

    2012-07-01

    Purpose: To evaluate the long-term toxicity, cosmesis, and local control of accelerated partial breast irradiation with implant brachytherapy after wide local excision for Stage T1N0 breast cancer (BCa). Materials and Methods: Between 1997 and 2001, 50 patients with Stage T1N0M0 BCa were treated in a Phase I-II protocol using low-dose-rate accelerated partial breast irradiation with implant brachytherapy after wide local excision and lymph node surgery. The total dose was escalated in three groups: 50 Gy (n = 20), 55 Gy (n = 17), and 60 Gy (n = 13). Patient- and physician-assessed breast cosmesis, patient satisfaction, toxicity, mammographic abnormalities, repeat biopsies, and disease status were prospectively evaluated at each visit. Kendall's tau ({tau}{sub {beta}}) and logistic regression analyses were used to correlate outcomes with dose, implant volume, patient age, and systemic therapy. Results: The median follow-up period was 11.2 years (range, 4-14). The patient satisfaction rate was 67%, 67% reported good-excellent cosmesis, and 54% had moderate-severe fibrosis. Higher dose was correlated with worse cosmetic outcome ({tau}{sub {beta}} 0.6, p < .0001), lower patient satisfaction ({tau}{sub {beta}} 0.5, p < .001), and worse fibrosis ({tau}{sub {beta}} 0.4, p = .0024). Of the 50 patients, 35% had fat necrosis and 34% developed telangiectasias {>=}1 cm{sup 2}. Grade 3-4 late skin and subcutaneous toxicities were seen in 4 patients (9%) and 6 patients (13%), respectively, and both correlated with higher dose ({tau}{sub {beta}} 0.3-0.5, p {<=} .01). One patient had Grade 4 skin ulceration and fat necrosis requiring surgery. Mammographic abnormalities were seen in 32% of the patients, and 30% underwent repeat biopsy, of which 73% were benign. Six patients had ipsilateral breast recurrence: five elsewhere in the breast, and one at the implant site. One patient died of metastatic BCa after recurrence. The 12-year actuarial local control, recurrence

  1. Accelerated Partial Breast Irradiation With Low-Dose-Rate Interstitial Implant Brachytherapy After Wide Local Excision: 12-Year Outcomes From a Prospective Trial

    International Nuclear Information System (INIS)

    Hattangadi, Jona A.; Powell, Simon N.; MacDonald, Shannon M.; Mauceri, Thomas; Ancukiewicz, Marek; Freer, Phoebe; Lawenda, Brian; Alm El-Din, Mohamed A.; Gadd, Michele A.; Smith, Barbara L.; Taghian, Alphonse G.

    2012-01-01

    Purpose: To evaluate the long-term toxicity, cosmesis, and local control of accelerated partial breast irradiation with implant brachytherapy after wide local excision for Stage T1N0 breast cancer (BCa). Materials and Methods: Between 1997 and 2001, 50 patients with Stage T1N0M0 BCa were treated in a Phase I-II protocol using low-dose-rate accelerated partial breast irradiation with implant brachytherapy after wide local excision and lymph node surgery. The total dose was escalated in three groups: 50 Gy (n = 20), 55 Gy (n = 17), and 60 Gy (n = 13). Patient- and physician-assessed breast cosmesis, patient satisfaction, toxicity, mammographic abnormalities, repeat biopsies, and disease status were prospectively evaluated at each visit. Kendall’s tau (τ β ) and logistic regression analyses were used to correlate outcomes with dose, implant volume, patient age, and systemic therapy. Results: The median follow-up period was 11.2 years (range, 4–14). The patient satisfaction rate was 67%, 67% reported good-excellent cosmesis, and 54% had moderate-severe fibrosis. Higher dose was correlated with worse cosmetic outcome (τ β 0.6, p β 0.5, p β 0.4, p = .0024). Of the 50 patients, 35% had fat necrosis and 34% developed telangiectasias ≥1 cm 2 . Grade 3–4 late skin and subcutaneous toxicities were seen in 4 patients (9%) and 6 patients (13%), respectively, and both correlated with higher dose (τ β 0.3–0.5, p ≤ .01). One patient had Grade 4 skin ulceration and fat necrosis requiring surgery. Mammographic abnormalities were seen in 32% of the patients, and 30% underwent repeat biopsy, of which 73% were benign. Six patients had ipsilateral breast recurrence: five elsewhere in the breast, and one at the implant site. One patient died of metastatic BCa after recurrence. The 12-year actuarial local control, recurrence-free survival, and overall survival rate was 85% (95% confidence interval, 70–97%), 72% (95% confidence interval, 54–86%), and 87% (95

  2. Endoscope-guided interstitial intensity-modulated brachytherapy and intracavitary brachytherapy as boost radiation for primary early T stage nasopharyngeal carcinoma.

    Directory of Open Access Journals (Sweden)

    Xiang-Bo Wan

    Full Text Available BACKGROUND: Intracavitary brachytherapy (ICBT is usually applied as boost radiotherapy for superficial residual of nasopharyngeal carcinoma (NPC after primary extern-beam radiptherapy (ERT. Here, we evaluated the outcome of endoscope-guided interstitial intensity-modulated brachytherapy (IMBT boost radiation for deep-seated residual NPC. METHODOLOGY/PRINCIPAL FINDINGS: Two hundred and thirteen patients with residual NPC who were salvaged with brachytherapy boost radiation during 2005-2009 were analyzed retrospectively. Among these patients, 171 patients had superficial residual NPC (≤1 cm below the nasopharyngeal epithelium were treated with ICBT boost radiation, and interstitial IMBT boost radiation was delivered to 42 patients with deep-seated residual NPC (>1 cm below the nasopharyngeal epithelium. We found that IMBT boost subgroup had a higher ratio of T2b (81.0% VS 34.5%, P<0.001 and stage II (90.5% VS 61.4%, P = 0.001 than that of ICBT boost subgroup. The dosage of external-beam radiotherapy in the nasopharyngeal (63.0±3.8 VS 62.6±4.3 Gray (Gy, P = 0.67 and regional lymph nodes (55.8±5.0 VS 57.5±5.7 Gy, P = 0.11 was comparable in both groups. For brachytherapy, IMBT subgroup had a lower boost radiation dosage than ICBT subgroup (11.0±2.9 VS 14.8±3.2 Gy, P<0.01. Though the IMBT group had deeper residual tumors and received lower boost radiation dosages, both subgroups had the similar 5-year actuarial overall survival rate (IMBT VS ICBT group: 96.8% VS 93.6%, P = 0.87, progression-free survival rate (92.4% VS 86.5%, P = 0.41 and distant metastasis-free survival rate (94.9% VS 92.7%, P = 0.64. Moreover, IMBT boost radiation subgroup had a similar local (97.4% VS 94.4%, P = 0.57 and regional (95.0% VS 97.2%, P = 0.34 control to ICBT subgroup. The acute and late toxicities rates were comparable between the both subgroups. CONCLUSIONS/SIGNIFICANCE: IMBT boost radiation may be a promising therapeutic

  3. Late complications after high-dose-rate interstitial brachytherapy for tongue cancer

    International Nuclear Information System (INIS)

    Shimizutani, Kimishige; Inoue, Takehiro; Inoue, Toshihiko; Yoshioka, Yasuo; Teshima, Teruki; Kakimoto, Naoya; Murakami, Shumei; Furukawa, Souhei; Fuchihata, Hajime

    2005-01-01

    The objectives of this study was to analyze the treatment results and late complications of high-dose-rate (HDR) interstitial brachytherapy (ISBT) for early (T1N0, T2N0) mobile tongue cancer using the microSelectron-HDR. From January 1993 through April 2001, a total of 72 patients with early squamous cell carcinomas of the mobile tongue were treated with microSelectron-HDR interstitial brachytherapy at the Department of Radiology, Osaka University Hospital. Of the patients, 18% were treated with a combination of prior external radiation and HDR-ISBT, and 82% were treated with HDR-ISBT alone. For HDR-ISBT alone, all cases were treated with a total dose of 54 Gy/9 fractions every 5 days or 60 Gy/10 fractions every 8 days. In combined therapy with an external dose of 30 to 40 Gy, HDR-ISBT was given at a total dose of 42-50 Gy. The Brinkman and alcohol indexes were used to analyze the incidence of late complications after HDR-ISBT. The 2- and 5-year local control rates were 85% and 82%, respectively. Fifteen of 72 patients (21%) treated with HDR-ISBT had late complications. Ten of 15 patients (67%) with late complications had a Brinkman index exceeding 600. HDR-ISBT is useful and easily applied under local anesthesia to early or superficial lesions of the mobile tongue. However, we found an increase in late complications, such as soft-tissue ulcers and bone exposure, after irradiation of tongue cancer with 60 Gy HDR-ISBT in patients with a Brinkman index greater than 600. (author)

  4. Evaluation of Geometrically Optimized Single- and Double-plane Interstitial High Dose Rate Implants with Respect to Conformality and Homogeneity

    International Nuclear Information System (INIS)

    Major, Tibor; Polgar, Csaba; Fodor, Janos; Takacsi-nagy, Zoltan; Mangel, Laszlo; Nemeth, Gyoergy

    2003-01-01

    The use of a stepping source in high dose rate brachytherapy supported with dwell-time optimization makes it possible to deviate from the classical dosimetry systems. Dose distributions of single- and double-plane implants were analysed for conformality and homogeneity at idealized target volumes. The Paris system was used for catheter positioning and target volume determination. Geometric optimization and individual dose prescription were applied. Volumetric indices and dose parameters were calculated at optimal active length, which was found to be equal to target volume length. The mean conformality, homogeneity, external volume and overdose volume indices were 0.78, 0.67, 0.22 and 0.13, respectively. The average minimum target and reference doses were 69% and 86%, respectively. Comparisons between the volumetric indices of geometrical optimized and non-optimized implants were also performed, and a significant difference was found regarding any index. The geometrical optimization resulted in superior conformality and slightly inferior homogeneity. At geometrically optimized implants, the active length can be reduced compared to non-optimized implants. Volumetric parameters and dose-volume histogram-based individual dose prescription are recommended for quantitative assessment of interstitial implants

  5. SU-E-T-507: Interfractional Variation of Fiducial Marker Position During HDR Brachytherapy with Cervical Interstitial Needle Template

    Energy Technology Data Exchange (ETDEWEB)

    Shen, S; Kim, R; Benhabib, S; Araujo, J; Burnett, L; Duan, J; Popple, R; Wu, X; Cardan, R; Brezovich, I [UniversityAlabama Birmingham, Birmingham, AL (United Kingdom)

    2014-06-01

    Purpose: HDR brachytherapy using interstitial needle template for cervical cancer is commonly delivered in 4-5 fractions. Routine verification of needle positions before each fraction is often based on radiographic imaging of implanted fiducial markers. The current study evaluated interfractional displacement of implanted fiducial markers using CT images. Methods: 9 sequential patients with cervical interstitial needle implants were evaluated. The superior and inferior borders of the target volumes were defined by fiducial markers in planning CT. The implant position was verified with kV orthogonal images before each fraction. A second CT was acquired prior 3rd fraction (one or 2 days post planning CT). Distances from inferior and superior fiducial markers to pubic symphysis plane (perpendicular to vaginal obtulator)were measured. Distance from needle tip of a reference needle (next to the inferior marker) to the pubic symphysis plane was also determined. The difference in fiducial marker distance or needle tip distance between planning CT and CT prior 3rd fraction were measured to assess markers migration and needle displacement. Results: The mean inferior marker displacement was 4.5 mm and ranged 0.9 to 11.3 mm. The mean superior marker displacement was 2.7 mm and ranged 0 to 10.4 mm. There was a good association between inferior and superior marker displacement (r=0.95). Mean averaged inferior and superior marker displacement was 3.3 mm and ranged from 0.1 to 10.9 mm, with a standard deviation of 3.2 mm. The mean needle displacement was 5.6 mm and ranged 0.2 to 15.6 mm. Needle displacements were reduced (p<0.05) after adjusting according to needle-to-fiducials distance. Conclusion: There were small fiducial marker displacements between HDR fractions. Our study suggests a target margin of 9.7 mm to cover interfractional marker displacements (in 95% cases) for pretreatment verification based on radiographic imaging.

  6. Interstitial rotating shield brachytherapy for prostate cancer

    International Nuclear Information System (INIS)

    Adams, Quentin E.; Xu, Jinghzu; Breitbach, Elizabeth K.; Li, Xing; Rockey, William R.; Kim, Yusung; Wu, Xiaodong; Flynn, Ryan T.; Enger, Shirin A.

    2014-01-01

    Purpose: To present a novel needle, catheter, and radiation source system for interstitial rotating shield brachytherapy (I-RSBT) of the prostate. I-RSBT is a promising technique for reducing urethra, rectum, and bladder dose relative to conventional interstitial high-dose-rate brachytherapy (HDR-BT). Methods: A wire-mounted 62 GBq 153 Gd source is proposed with an encapsulated diameter of 0.59 mm, active diameter of 0.44 mm, and active length of 10 mm. A concept model I-RSBT needle/catheter pair was constructed using concentric 50 and 75 μm thick nickel-titanium alloy (nitinol) tubes. The needle is 16-gauge (1.651 mm) in outer diameter and the catheter contains a 535 μm thick platinum shield. I-RSBT and conventional HDR-BT treatment plans for a prostate cancer patient were generated based on Monte Carlo dose calculations. In order to minimize urethral dose, urethral dose gradient volumes within 0–5 mm of the urethra surface were allowed to receive doses less than the prescribed dose of 100%. Results: The platinum shield reduced the dose rate on the shielded side of the source at 1 cm off-axis to 6.4% of the dose rate on the unshielded side. For the case considered, for the same minimum dose to the hottest 98% of the clinical target volume (D 98% ), I-RSBT reduced urethral D 0.1cc below that of conventional HDR-BT by 29%, 33%, 38%, and 44% for urethral dose gradient volumes within 0, 1, 3, and 5 mm of the urethra surface, respectively. Percentages are expressed relative to the prescription dose of 100%. For the case considered, for the same urethral dose gradient volumes, rectum D 1cc was reduced by 7%, 6%, 6%, and 6%, respectively, and bladder D 1cc was reduced by 4%, 5%, 5%, and 6%, respectively. Treatment time to deliver 20 Gy with I-RSBT was 154 min with ten 62 GBq 153 Gd sources. Conclusions: For the case considered, the proposed 153 Gd-based I-RSBT system has the potential to lower the urethral dose relative to HDR-BT by 29%–44% if the clinician allows

  7. Effect of brachytherapy technique and patient characteristics on cervical cancer implant dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Anker, Christopher J., E-mail: chris.anker@hci.utah.edu [Department of Radiation Oncology, Huntsman Cancer Hospital, University of Utah, Salt Lake City, UT (United States); O' Donnell, Kristen [Department of Radiation Oncology, The University of Arizona, Tucson, AZ (United States); Boucher, Kenneth M. [Department of Oncological Sciences, Huntsman Cancer Institute, University of Utah, Salt Lake City, UT (United States); Gaffney, David K. [Department of Radiation Oncology, Huntsman Cancer Hospital, University of Utah, Salt Lake City, UT (United States)

    2013-01-01

    Our purpose was to evaluate the relationship between brachytherapy technique and patient characteristics on dose to organs-at-risk (OARs) in patients undergoing high dose rate (HDR) brachytherapy for cervical cancer. From 1998 to 2008, 31 patients with cervical cancer with full dosimetric data were identified who received definitive external-beam radiation and HDR brachytherapy with tandem and ovoid applicators. Doses were recorded at point A, the International Commission on Radiation Units and Measurements (ICRU)-38 rectal point, the ICRU-38 bladder point, the vaginal surface, and the pelvic sidewall. Generalized estimating equations were used to determine the significance of changes in OAR to point A dose ratios with differences in brachytherapy technique or patient characteristics. Patients underwent a median of 5 brachytherapy procedures (range, 3 to 5), with a total of 179 procedures for 31 patients. For all brachytherapy treatments, the average ratios between the doses for the rectal, bladder, vaginal surface, and pelvic sidewall reference points to those at point A were 0.49, 0.59, 1.15, and 0.17, respectively. In general, decreased OAR dose was associated with a lower stage, younger age, increased ovoid size, increased tandem length, and earlier implant number. Increased tandem curvature significantly increased bladder dose and decreased rectal dose. Intravenous anesthesia usage was not correlated with improved dosimetry. This study allowed identification of patient and procedure characteristics influencing OAR dosing. Although the advent of 3-dimensional (3D) image-guided brachytherapy will bring new advances in treatment optimization, the actual technique involved at the time of the brachytherapy implant procedure will remain important.

  8. Rectal complications associated with transperineal interstitial brachytherapy for prostate cancer

    International Nuclear Information System (INIS)

    Gelblum, Daphna Y.; Potters, Louis

    2000-01-01

    Purpose: As transperineal interstitial permanent prostate brachytherapy (TIPPB) grows in acceptance as an option in the treatment of organ-confined prostate cancer, its associated toxicities are being defined. This clinical report documents rectal toxicity from a large cohort of men treated by a single practitioner for adenocarcinoma of the prostate. Methods and Materials: Eight hundred twenty-five men were treated from September 1992 to September 1998 with TIPPB. One hundred-forty were treated in conjunction with external beam irradiation (EBRT) and 685 with TIPPB alone. All patients were implanted under real-time ultrasound guidance. No dose-volume histogram analysis was performed for this study. All patients were followed at 5 weeks after the procedure, then every 3-6 months thereafter. Rectal morbidity was graded by a modified RTOG toxicity scale. Therapy to control symptoms was recommended on an individual basis. Results: The median follow-up for the cohort is 48 months. A total of 77 patients (9.4%) reported Grade 1 toxicity at some time following an implant whereas 54 patients (6.6%) reported Grade 2 toxicity. The peak post-TIPPB time for experiencing rectal toxicity was 8 months at which time Grade 1 and 2 rectal toxicity was reported in 9.5% of the patients. This improved over the subsequent months and resolved in all patients by 3((1)/(2)) years. Four patients (0.5%) reported Grade 3 rectal toxicity with rectal ulceration identified on colonoscopy at 1 year from implant. Two of the four patients had colonic manipulation in the radiated portion of the colon which subsequently caused it to bleed. None of the patients required blood product transfusion. In 3 of the 4 patients the Grade 3 rectal toxicity has resolved spontaneously and 1 patient continues to heal at the time of this report. No patient required hospitalization or surgical intervention. Conclusion: TIPPB is a tolerable and acceptable treatment option when used alone in early-stage, organ

  9. The effectiveness of 125I seed interstitial brachytherapy for transplantation tumor of human pancreatic carcinoma in nude mice: an experiment in vivo

    International Nuclear Information System (INIS)

    Song Qi; Liu Yu; Wang Zhongmin; Huang Wei; Lu Jian; Chen Kemin

    2010-01-01

    Objective: To discuss the effectiveness and therapeutic mechanism of 125 I interstitial brachytherapy for transplantation tumor of human pancreatic carcinoma in nude mice. Methods: The human pancreatic cell line Sw1990 was subcutaneously injected into the right lower limb partially dorsal area next to the groin of the immunodeficient BABL /c nude mice. The tumor was removed and cut into small pieces after it was formed,then the tumor pieces were inoculated in nude mice. The tumor developed to 8-10 mm in size after six weeks. A total of 16 nude mice with the suitable tumor size were used in this study. The 16 experimental mice were randomly and equally divided into two groups. The mice in study group (n = 8) were implanted with 125 I seeds, while the mice in control group (n = 8) were implanted with ghost seeds. After the implantation both the long and short diameter of the tumors as well as the mouse body weight were measured every 4 days. The tumor weight was measured when the mouse was sacrificed. The paraffin-embedded samples were sent for histopathological examination. Apoptotic cells were checked with terminal deoxynucleotidyl transferase dUTP nick end labeling (TUNEL) method. Expression of proliferating cell nuclear antigen (PCNA) was detected with immuno-histochemical staining. Results: The tumor grew slowly in the study group, but rapidly in the control group. The tumor weight in the study group and the control group was (2.68 ± 0.70)g and (4.68 ± 1.45)g, respectively, the difference between two groups was statistically significant (P = 0.021). The tumor inhibition rate was about 42.66%. No significant difference in body weight of nude mice existed between two groups both before and after the treatment (P > 0.05). Marked tumor necrosis was seen in study group, but no obvious, or only a little, tumor necrosis could be observed in the control group. The apoptotic index checked with the TUENL method in the study group and control group was (23.2 ± 1.9)% and

  10. Spectral CT evaluation of interstitial brachytherapy in pancreatic carcinoma xenografts: preliminary animal experience

    Energy Technology Data Exchange (ETDEWEB)

    Hu, Shudong [Jiangsu University, Department of Radiology, The Affiliated Renmin Hospital, Zhenjiang, Jiangsu (China); Shanghai Jiao tong University, School of Medicine, Department of Radiology, Ruijin Hospital, Shanghai (China); Huang, Wei; Song, Qi; Lin, Xiaozhu; Wang, Zhongmin; Chen, Kemin [Shanghai Jiao tong University, School of Medicine, Department of Radiology, Ruijin Hospital, Shanghai (China); Chen, Yerong [Jiangsu University, Department of Radiology, The Affiliated Renmin Hospital, Zhenjiang, Jiangsu (China)

    2014-09-15

    We sought to evaluate the capability of spectral CT to detect the therapeutic response to {sup 125}I interstitial brachytherapy in a pancreatic carcinoma xenograft nude mouse model. Twenty mice bearing SWl990 human pancreatic cancer cell xenografts were randomly separated into two groups: experimental (n = 10; 1.0 mCi) and control (n = 10; 0 mCi). After a two-week treatment, spectral CT was performed. Contrast-to-noise ratio (CNR) and iodine concentration (IC) in the lesions were measured and normalized to the muscle tissue, and nIC CD31 immunohistochemistry was used to measure microvessel density (MVD). The relationships between the nIC and MVD of the tumours were analysed. The nIC of the experimental group was significantly lower than that of the control group during the multiphase examination. A significant difference in the MVD was observed between the two groups (P <0.001). The nIC values of the three-phase scans have a certain positive correlation with MVD (r = 0.57, p < 0.0001; r = 0.48, p = 0.002; r = 0.63, p = 0.0017 in the 10, 25, and 60 s phase, respectively). Spectral CT can be a useful non-invasive imaging modality in evaluating the therapeutic effect of {sup 125}I interstitial brachytherapy to a pancreatic carcinoma. (orig.)

  11. Iodine-125 seed implantation (permanent brachytherapy) for clinically localized prostate cancer

    International Nuclear Information System (INIS)

    Ebara, Shin; Katayama, Yoshihisa; Tanimoto, Ryuta

    2008-01-01

    From January 2004 to March 2007, 308 patients with clinically localized prostate cancer were treated using iodine-125 ( 125 I) seed implantation (permanent brachytherapy) at Okayama University Graduate School of Medicine, Dentistry and Pharmaceutical Sciences. We evaluated the treatment's efficacy and morbidity in 300 prostate cancer patients who were followed up for more than 1 month after brachytherapy. Based on the National Comprehensive Cancer Network (NCCN) guidelines, patients with a prostate volume of less than 40 ml in transrectal ultrasound imaging were classified as low or intermediate risk. The median patient age was 67 years (range 50 to 79 years), the median prostate-specific antigen (PSA) value before biopsy was 6.95 ng/ml (range 1.13 to 24.7 ng/ml), and the median prostate volume was 24.33 ml (range 9.3 to 41.76 ml). The median follow-up was 18 months (range 1 to 36 months) and the PSA levels decreased in almost all patients after brachytherapy. Although 194 of 300 patients (64.7%) complained of difficulty in urination, pollakisuria/urgency, miction pain, and/or urinary incontinence, all of which might be associated with radiation prostatitis during the first month after brachytherapy, these symptoms gradually improved. 125 I seed implantation brachytherapy is safe and effective for localized prostate cancer within short-term follow up. (author)

  12. Volume correction factor in time dose relationships in brachytherapy

    International Nuclear Information System (INIS)

    Supe, S.J.; Sasane, J.B.

    1987-01-01

    Paterson's clinical data about the maximum tolerance doses for various volumes of interstitial implants with Ra-226 delivered in seven days was made use of in deriving volume correction factors for TDF and CRE concepts respectively for brachytherapy. The derived volume correction factors for TDF and for CRE differ fromthe one assumed for CRE by Kirk et al. and implied for TDF by Goitein. A normalising volume of 70 cc has been suggested for both CRE and TDF concepts for brachytherapy. A table showing the volume corrected TDF is presented for various volumes and dose rates for continuous irradiation. The use of this table is illustrated with examples. (orig.) [de

  13. Image and laparoscopic guided interstitial brachytherapy for locally advanced primary or recurrent gynaecological cancer using the adaptive GEC ESTRO target concept

    International Nuclear Information System (INIS)

    Fokdal, Lars; Tanderup, Kari; Nielsen, Soren Kynde; Christensen, Henrik Kidmose; Rohl, Lisbeth; Pedersen, Erik Morre; Schonemann, Niels Kim; Lindegaard, Jacob Christian

    2011-01-01

    Purpose: To retrospectively assess treatment outcome of image and laparoscopic guided interstitial pulsed dose rate brachytherapy (PDR-BT) for locally advanced gynaecological cancer using the adaptive GEC ESTRO target concept. Materials and methods: Between June 2005 and December 2010, 28 consecutive patients were treated for locally advanced primary vaginal (nine), recurrent endometrial (12) or recurrent cervical cancer (seven) with combined external beam radiotherapy (EBRT) and interstitial PDR-BT. Treatment was initiated with whole pelvic EBRT to a median dose of 45 Gy followed by PDR-BT using the Martinez Universal Perineal Interstitial Template (MUPIT). All implants were virtually preplanned using MRI of the pelvis with a dummy MUPIT in situ. The GEC ESTRO high risk clinical target volume (HR CTV), intermediate risk clinical target volume (IR CTV) and the organs at risk (OAR) were contoured and a preplan for implantation was generated (BrachyVision, Varian). The subsequent implantation was performed under laparoscopic visualisation. Final contouring and treatment planning were done using a post-implant CT. Planning aim of PDR-BT was to deliver 30 Gy in 50 hourly pulses to HR CTV. Manual dose optimisation was performed with the aim of reaching a D90 > 80 Gy in the HR CTV calculated as the total biologically equivalent to 2 Gy fractions of EBRT and BT (EQD2). Dose to the OAR were evaluated using dose volume constraints for D 2cc of 90 Gy for bladder and 70 Gy for rectum and sigmoid. Results: For HR CTV the median volume was 26 cm 3 (7-91 cm 3 ). Coverage of the HR CTV was 97% (90-100%) and D90 was 82 Gy (77-88 Gy). The D 2cc for bladder, rectum, and sigmoid were 65 Gy (47-81 Gy), 61 Gy (50-77 Gy), and 52 Gy (44-68 Gy), respectively. Median follow up was 18 months (6-61 months). The actuarial 2 years local control rate was 92% (SE 5), while disease-free survival and overall survival were 59% (SE 11) and 74%, respectively (SE 10). No complications to the

  14. Brachytherapy for Buccal Cancer: From Conventional Low Dose Rate (LDR) or Mold Technique to High Dose Rate Interstitial Brachytherapy (HDR-ISBT).

    Science.gov (United States)

    Kotsuma, Tadayuki; Yamazaki, Hideya; Masui, Koji; Yoshida, Ken; Shimizutani, Kimishige; Akiyama, Hironori; Murakami, Shumei; Isohashi, Fumiaki; Yoshioka, Yasuo; Ogawa, Kazuhiko; Tanaka, Eiichi

    2017-12-01

    To examine the effectiveness of newly-installed high-dose-rate interstitial brachytherapy (HDR-ISBT) for buccal cancer. We retrospectively reviewed 36 patients (25 men and 11 women) with buccal cancer treated with curative brachytherapy with or without external radiotherapy with a median follow-up of 99 months. A total of 15 HDR-ISBT (median 48 Gy/ 8 fractions, range=24-60 Gy) patients were compared to conventional 15 cases LDR-ISBT (70 Gy, range=42.8-110 Gy) and 7 molds techniques (15 Gy, range=9-74 Gy). A total of 31 patients also underwent external radiotherapy (30 Gy, range=24-48 Gy). They comprised of 3T1, 23 T2, 8 T3, 3 T4 including 11 node positive cases. HDR-ISBT provided 82% of local control rate at 5 years, whereas conventional brachytherapy showed 72% [p=0.44; LDR-ISBT (65%), mold therapy (85.7%)]. Patients with early lesions (T1-2 or stage I-II) showed better local control rates than those with advanced lesions (T3-4 or stage III-IV). Severe late grade 3 complications developed in two patients treated with LDR-ISBT and EBRT. There is no significant difference in toxicity grade ≤2 between conventional brachytherapy (5/15=33%) and HDR-ISBT (7/32=32%, p=0.92). HDR-ISBT achieved good and comparable local control rates to conventional brachytherapy without elevating the toxicity. Copyright© 2017, International Institute of Anticancer Research (Dr. George J. Delinasios), All rights reserved.

  15. Image guided adaptive brachytherapy with combined intracavitary and interstitial technique improves the therapeutic ratio in locally advanced cervical cancer: Analysis from the retroEMBRACE study

    DEFF Research Database (Denmark)

    LU, Fokdal; Sturdza, Alina; Mazeron, Renaud

    2016-01-01

    Background and purpose Image guided adaptive brachytherapy (IGABT) using intracavitary applicators (IC) has led to a significant improvement of local control in locally advanced cervical cancer (LACC). Further improvement has been obtained with combined intracavitary/interstitial (IC/IS) applicat...... IC/IS brachytherapy improves the therapeutic ratio in LACC by enabling a tumour specific dose escalation resulting in significantly higher local control in large tumours without adding treatment related late morbidity.......Background and purpose Image guided adaptive brachytherapy (IGABT) using intracavitary applicators (IC) has led to a significant improvement of local control in locally advanced cervical cancer (LACC). Further improvement has been obtained with combined intracavitary/interstitial (IC....../IS) applicators. The aim of this analysis was to evaluate the impact on local control and late morbidity of application of combined IS/IC brachytherapy in a large multicentre population. Material/methods 610 patients with LACC from the retroEMBRACE study were included. Patients were divided into an IC group (N...

  16. 3D-navigation for interstitial stereotactic brachytherapy; 3D-Navigation in der interstitiellen stereotaktischen Brachytherapie

    Energy Technology Data Exchange (ETDEWEB)

    Auer, T.; Hensler, E.; Eichberger, P.; Bluhm, A.; Lukas, P. [Innsbruck Univ. (Austria). Klinik fuer Strahlentherapie und Radioonkologie; Gunkel, A.; Freysinger, W.; Bale, R.; Thumfart, W.F. [Innsbruck Univ. (Austria). Klinik fuer HNO-Krankheiten; Gaber, O. [Innsbruck Univ. (Austria). Inst. fuer Anatomie

    1998-02-01

    The aim of this paper is to describe the adaption of 3D-navigation for interstitial brachytherapy. The new method leads to prospective and therefore improved planning of the therapy (position of the needle and dose distribution) and to the possibility of a virtual simulation (control if vessels or nerves are on the pathway of the needle). The EasyGuide Neuro {sup trademark} navigation system (Philips) was adapted in the way, that needles for interstitial bracachytherapy were made connectable to the pointer and correctly displayed on the screen. To determine the positioning accuracy, several attempts were performed to hit defined targets on phantoms. Two methods were used: `Free navigation`, where the needle was under control of the navigation system, and the `guided navigation` where an aligned template was used additionally to lead the needle to the target. In addition a mask system was tested, whether it met the requirements of stable and reproducible positioning. The potential of applying this method is clinical practice was tested with an anatomical specimen. About 91% of all attempts lied within 5 mm. There were even better results on the more rigid table (94%<4 mm). No difference could be seen between both application methods (`free navigation` and `navigation with template`), they showed the same accuracy. (orig./MG) [Deutsch] Es war das Ziel dieser Arbeit, ein 3D-Infrarotnavigationssystem fuer die Anforderungen der interstitiellen stereotaktischen Brachytherapie zu adaptieren. Damit wird die Planung der Therapie verbessert (prospektive Planung der Nadelpositionen und der Dosisverteilung), und eine virtuelle Simulation wird realisierbar (Kontrolle des vorgeplanten Zugangs bezueglich Verletzungsmoeglichkeit von Gefaessen oder Nerven). Das EasyGuide-Neuro {sup trademark} -Navigagationssystem (Philips) wurde so veraendert, dass Nadeln, die in der Brachytherapie Verwendung finden, am Pointer befestigt werden konnten und am Bildschirm angezeigt wurden. Um die

  17. Interstitial brachytherapy for liver metastases and assessment of response by positron emission tomography: a case report

    Directory of Open Access Journals (Sweden)

    Goura Kishor Rath

    2010-10-01

    Full Text Available For liver metastases (LM, image guided percutaneous ablative procedures such as radiofrequency ablation (RFA, laser induced thermal therapy (LITT and trans-arterial chemo-embolisation (TACE are increasingly being used because they are relatively safer, less invasive and equally effective. CT scan guided interstitial brachytherapy (IBT with a single large dose of radiation by high dose rate (HDR brachytherapy is a novel technique of treating LM and has shown good results. Positron emission tomography (PET scan may provide better information for assessing the response toIBT procedures. We hereby report a case of LM that was treated by HDR IBT and PET scan was done in addition to CT scan for assessing the response.

  18. Brachytherapy in head and neck cancers; Curietherapie des cancers de la sphere ORL

    Energy Technology Data Exchange (ETDEWEB)

    Mazeron, J.J.; Noel, G.; Simon, J.M.; Racadot, S.; Jauffret, E. [Groupe Hospitalier la Pitie-Salpetriere, Centre des Tumeurs, 75 - Paris (France)

    2003-02-01

    Experience accumulated over several decades with radiation of Head and Neck tumours by irradiation has demonstrated the need for a high tumour dose to achieve local control. With external beam irradiation alone, it is difficult to spare adjacent normal tissues, resulting in undesirable late effects on the salivary glands; mandible, and muscles of mastication. Interstitial implantation is ideally suited to deliver a high dose limited to the volume of the primary tumor, thus minimizing sequels. A large experience has been accumulated with low dose rate (LDR) brachytherapy in treatment of carcinoma of oral cavity, oropharynx, and nasopharynx. Recent analysis of large clinical series provided data indicating that modalities of low dose rate brachytherapy should be optimized in treating these tumors for increasing therapeutic ratio. Low dose rate brachytherapy is now challenged by high dose rate (HDR) brachytherapy and pulsed dose rate (PDR) brachytherapy. Preliminary results obtained with these two last modalities are discussed regarding to those of low dose rate brachytherapy. (authors)

  19. Salvage prostate HDR brachytherapy combined with interstitial hyperthermia for local recurrence after radiation therapy failure

    Energy Technology Data Exchange (ETDEWEB)

    Kukielka, A.M.; Hetnal, M.; Dabrowski, T.; Walasek, T.; Brandys, P.; Reinfuss, M. [Centre of Oncology, M. Sklodowska - Curie Institute, Krakow Branch, Department of Radiotherapy, Krakow (Poland); Nahajowski, D.; Kudzia, R.; Dybek, D. [Centre of Oncology, M. Sklodowska - Curie Institute, Krakow Branch, Department of Medical Physics, Department of Radiotherapy, Krakow (Poland)

    2014-02-15

    The aim of the present retrospective study is to evaluate toxicity and early clinical outcomes of interstitial hyperthermia (IHT) combined with high-dose rate (HDR) brachytherapy as a salvage treatment in patients with biopsy-confirmed local recurrence of prostate cancer after previous external beam radiotherapy. Between September 2008 and March 2013, 25 patients with local recurrence of previously irradiated prostate cancer were treated. The main eligibility criteria for salvage prostate HDR brachytherapy combined with interstitial hyperthermia were biopsy confirmed local recurrence and absence of nodal and distant metastases. All patients were treated with a dose of 30 Gy in 3 fractions at 21-day intervals. We performed 62 hyperthermia procedures out of 75 planned (83 %). The aim of the hyperthermia treatment was to heat the prostate to 41-43 C for 60 min. Toxicity for the organs of the genitourinary system and rectum was assessed according to the Common Terminology Criteria for Adverse Events (CTCAE, v. 4.03). Determination of subsequent biochemical failure was based on the Phoenix definition (nadir + 2 ng/ml). The median age was 71 years (range 62-83 years), the median initial PSA level was 16.3 ng/ml (range 6.37-64 ng/ml), and the median salvage PSA level was 2.8 ng/ml (1.044-25.346 ng/ml). The median follow-up was 13 months (range 4-48 months). The combination of HDR brachytherapy and IHT was well tolerated. The most frequent complications were nocturia, weak urine stream, urinary frequency, hematuria, and urgency. Grade 2 rectal hemorrhage was observed in 1 patient. No grade 3 or higher complications were observed. The 2-year Kaplan-Meier estimate of biochemical control after salvage treatment was 74 %. The PSA in 20 patients decreased below the presalvage level, while 11 patients achieved a PSA nadir < 0.5 ng/ml. All patients are still alive. Of the 7 patients who experienced biochemical failure, bone metastases were found in 2 patients. IHT in combination

  20. Interstitial high-dose-rate brachytherapy boost: The feasibility and cosmetic outcome of a fractionated outpatient delivery scheme

    International Nuclear Information System (INIS)

    Manning, Matthew A.; Arthur, Douglas W.; Schmidt-Ullrich, Rupert K.; Arnfield, Mark R.; Amir, Cyrus; Zwicker, Robert D.

    2000-01-01

    Purpose: To evaluate the feasibility, potential toxicity, and cosmetic outcome of fractionated interstitial high dose rate (HDR) brachytherapy boost for the management of patients with breast cancer at increased risk for local recurrence. Methods and Materials: From 1994 to 1996, 18 women with early stage breast cancer underwent conventionally fractionated whole breast radiotherapy (50-50.4 Gy) followed by interstitial HDR brachytherapy boost. All were considered to be at high risk for local failure. Seventeen had pathologically confirmed final surgical margins of less than 2 mm or focally positive. Brachytherapy catheter placement and treatment delivery were conducted on an outpatient basis. Preplanning was used to determine optimal catheter positions to enhance dose homogeneity of dose delivery. The total HDR boost dose was 15 Gy delivered in 6 fractions of 2.5 Gy over 3 days. Local control, survival, late toxicities (LENT-SOMA), and cosmetic outcome were recorded in follow-up. In addition, factors potentially influencing cosmesis were analyzed by logistic regression analysis. Results: The minimum follow-up is 40 months with a median 50 months. Sixteen patients were alive without disease at last follow-up. There have been no in-breast failures observed. One patient died with brain metastases, and another died of unrelated causes without evidence of disease. Grade 1-2 late toxicities included 39% with hyperpigmentation, 56% with detectable fibrosis, 28% with occasional discomfort, and 11% with visible telangiectasias. Grade 3 toxicity was reported in one patient as persistent discomfort. Sixty-seven percent of patients were considered to have experienced good/excellent cosmetic outcomes. Factors with a direct relationship to adverse cosmetic outcome were extent of surgical defect (p = 0.00001), primary excision volume (p = 0.017), and total excision volume (p = 0.015). Conclusions: For high risk patients who may benefit from increased doses, interstitial HDR

  1. Interfraction patient motion and implant displacement in prostate high dose rate brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Fox, C. D.; Kron, T.; Leahy, M.; Duchesne, G.; Williams, S.; Tai, K. H.; Haworth, A.; Herschtal, A.; Foroudi, F. [Department of Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Victoria 3002 (Australia); Nursing Service, Peter MacCallum Cancer Centre, Melbourne, Victoria 3002 (Australia); Department of Radiation Oncology, Peter MacCallum Cancer Centre and University of Melbourne, Melbourne, Victoria 3002 (Australia); Department of Physical Sciences, Peter MacCallum Cancer Centre and Royal Melbourne Insititute of Technology, Melbourne, Victoria 3000 (Australia); Department of Biostatistics and Clinical Trials, Peter MacCallum Cancer Centre, Melbourne, Victoria 3002 (Australia); Department of Radiation Oncology, Peter MacCallum Cancer Centre and University of Melbourne, Melbourne, Victoria 3010 (Australia)

    2011-11-15

    Purpose: To quantify movement of prostate cancer patients undergoing treatment, using an in-house developed motion sensor in order to determine a relationship between patient movement and high dose rate (HDR) brachytherapy implant displacement. Methods: An electronic motion sensor was developed based on a three axis accelerometer. HDR brachytherapy treatment for prostate is delivered at this institution in two fractions 24 h apart and 22 patients were monitored for movement over the interval between fractions. The motion sensors functioned as inclinometers, monitoring inclination of both thighs, and the inclination and roll of the abdomen. The implanted HDR brachytherapy catheter set was assessed for displacement relative to fiducial markers in the prostate. Angle measurements and angle differences over a 2 s time base were binned, and the standard deviations of the resulting frequency distributions used as a metric for patient motion in each monitored axis. These parameters were correlated to measured catheter displacement using regression modeling. Results: The mean implant displacement was 12.6 mm in the caudal direction. A mean of 19.95 h data was recorded for the patient cohort. Patients generally moved through a limited range of angles with a mean of the exception of two patients who spent in excess of 2 h lying on their side. When tested for a relationship between movement in any of the four monitored axes and the implant displacement, none was significant. Conclusions: It is not likely that patient movement influences HDR prostate implant displacement. There may be benefits to patient comfort if nursing protocols were relaxed to allow patients greater freedom to move while the implant is in situ.

  2. Interfraction patient motion and implant displacement in prostate high dose rate brachytherapy

    International Nuclear Information System (INIS)

    Fox, C. D.; Kron, T.; Leahy, M.; Duchesne, G.; Williams, S.; Tai, K. H.; Haworth, A.; Herschtal, A.; Foroudi, F.

    2011-01-01

    Purpose: To quantify movement of prostate cancer patients undergoing treatment, using an in-house developed motion sensor in order to determine a relationship between patient movement and high dose rate (HDR) brachytherapy implant displacement. Methods: An electronic motion sensor was developed based on a three axis accelerometer. HDR brachytherapy treatment for prostate is delivered at this institution in two fractions 24 h apart and 22 patients were monitored for movement over the interval between fractions. The motion sensors functioned as inclinometers, monitoring inclination of both thighs, and the inclination and roll of the abdomen. The implanted HDR brachytherapy catheter set was assessed for displacement relative to fiducial markers in the prostate. Angle measurements and angle differences over a 2 s time base were binned, and the standard deviations of the resulting frequency distributions used as a metric for patient motion in each monitored axis. These parameters were correlated to measured catheter displacement using regression modeling. Results: The mean implant displacement was 12.6 mm in the caudal direction. A mean of 19.95 h data was recorded for the patient cohort. Patients generally moved through a limited range of angles with a mean of the exception of two patients who spent in excess of 2 h lying on their side. When tested for a relationship between movement in any of the four monitored axes and the implant displacement, none was significant. Conclusions: It is not likely that patient movement influences HDR prostate implant displacement. There may be benefits to patient comfort if nursing protocols were relaxed to allow patients greater freedom to move while the implant is in situ.

  3. Low-dose-rate interstitial brachytherapy preserves good quality of life in buccal mucosa cancer patients

    International Nuclear Information System (INIS)

    Tayier, A.; Hayashi, Keiji; Yoshimura, Ryoichi

    2011-01-01

    The purpose of this study was to determine the results and long-term changes in radiation toxicity of stage I-II buccal mucosa cancer patients treated by low-dose-rate (LDR) brachytherapy with 198 Au grains. A total of 133 stage I-II buccal mucosa carcinomas patients received 198 Au grain implantation brachytherapy between January 1982 and July 2005: 75 of them were treated by 198 Au grain implantation alone and 58 were treated by 198 Au implantation in combination with external irradiation. The average 198 Au-grain dose was 70 Gy in 7 days. Gross tumor areas ranged from 2.4 cm 2 to 9 cm 2 , and the clinical target areas ranged from 6 cm 2 to 15 cm 2 . The follow-up periods ranged from 3 months to 20 years (mean: 5 years 11 months and median: 5 years 1 months). Failure at the site of the primary lesion occurred in 17 patients. Post-treatment mucosal ulceration developed in 15 patients, and all were cured within 25 months by conservative treatment. Osteoradionecrosis was diagnosed in 8 patients, but only one patient required surgical treatment. No severe complications or aggravation of complications developed more than 10 years after treatment. The results of low-dose-rate (LDR)-brachytherapy (BT) alone and LDR-BT in combination with external irradiation at a total dose of 25 Gy were acceptable from the standpoint of cure rate and quality of life (QOL). (author)

  4. Accelerated partial-breast irradiation using high-dose-rate interstitial brachytherapy: 12-year update of a prospective clinical study

    International Nuclear Information System (INIS)

    Polgar, Csaba; Major, Tibor; Fodor, Janos; Sulyok, Zoltan; Somogyi, Andras; Loevey, Katalin; Nemeth, Gyoergy; Kasler, Miklos

    2010-01-01

    Background and purpose: To report the 12-year updated results of accelerated partial-breast irradiation (APBI) using multicatheter interstitial high-dose-rate (HDR) brachytherapy (BT). Patients and methods: Forty-five prospectively selected patients with T1N0-N1mi, nonlobular breast cancer without the presence of an extensive intraductal component and with negative surgical margins were treated with APBI after breast-conserving surgery (BCS) using interstitial HDR BT. A total dose of 30.3 Gy (n = 8) and 36.4 Gy (n = 37) in seven fractions within 4 days was delivered to the tumour bed plus a 1-2 cm margin. The median follow-up time was 133 months for surviving patients. Local and regional control, disease-free (DFS), cancer-specific (CSS), and overall survival (OS), as well as late side effects, and cosmetic results were assessed. Results: Four (8.9%) ipsilateral breast tumour recurrences were observed, for a 5-, 10-, and 12-year actuarial rate of 4.4%, 9.3%, and 9.3%, respectively. A total of two regional nodal failures were observed for a 12-year actuarial rate of 4.4%. The 12-year DFS, CSS, and OS was 75.3%, 91.1%, and 88.9%, respectively. Grade 3 fibrosis was observed in one patient (2.2%). No patient developed grade 3 teleangiectasia. Fat necrosis requiring surgical intervention occurred in one woman (2.2%). Cosmetic results were rated excellent or good in 35 patients (77.8%). Conclusions: Twelve-year results with APBI using HDR multicatheter interstitial implants continue to demonstrate excellent long-term local tumour control, survival, and cosmetic results with a low-rate of late side effects.

  5. Feasibility of salvage interstitial microwave thermal therapy for prostate carcinoma following failed brachytherapy: studies in a tissue equivalent phantom

    International Nuclear Information System (INIS)

    McCann, Claire; Kumaradas, J Carl; Gertner, Mark R; Davidson, Sean R H; Dolan, Alfred M; Sherar, Michael D

    2003-01-01

    Thermal therapy is an experimental treatment to destroy solid tumours by heating them to temperatures ranging from 55 deg C to 90 deg C, inducing thermal coagulation and necrosis of the tumour. We are investigating the feasibility of interstitial microwave thermal therapy as a salvage treatment for prostate cancer patients with local recurrence following failed brachytherapy. Due to the electrical and thermal conductivity of the brachytherapy seeds, we hypothesized that the seeds could scatter the microwave energy and cause unpredictable heating. To investigate this, a 915 MHz helical antenna was inserted into a muscle-equivalent phantom with and without brachytherapy seeds. Following a 10 W, 5 s input to the antenna, the temperature rise was used to calculate absorbed power, also referred to as specific absorption rate (SAR). Plane wave models based on Maxwell's equations were also used to characterize the electromagnetic scattering effect of the seeds. In addition, the phantom was heated with 8 W for 5 min to quantify the effect of the seeds on the temperature distribution during extended heating. SAR measurements indicated that the seeds had no significant effect on the shape and size of the SAR pattern of the antenna. However, the plane wave simulations indicated that the seeds could scatter the microwave energy resulting in hot spots at the seed edges. Lack of experimental evidence of these hot spots was probably due to the complex polarization of the microwaves emitted by the helical antenna. Extended heating experiments also demonstrated that the seeds had no significant effect on the temperature distributions and rates of temperature rise measured in the phantom. The results indicate that brachytherapy seeds are not a technical impediment to interstitial microwave thermal therapy as a salvage treatment following failed brachytherapy

  6. Dosimetric intercomparison of permanent Ho-166 seed's implants and HDR Ir-192 brachytherapy in breast cancer.

    Science.gov (United States)

    de Campos, Tarcisio Passos Ribeiro; Nogueira, Luciana Batista; Trindade, Bruno; Cuperschmid, Ethel Mizrahy

    2016-01-01

    To provide a comparative dosimetric analysis of permanent implants of Ho(166)-seeds and temporary HDR Ir(192)-brachytherapy through computational simulation. Brachytherapy with Ir(192)-HDR or LDR based on temporary wires or permanent radioactive seed implants can be used as dose reinforcement for breast radiation therapy. Permanent breast implants have not been a practical clinical routine; although, I(125) and Pd(103)-seeds have already been reported. Biodegradable Ho(166)-ceramic-seeds have been addressed recently. Simulations of implants of nine Ho(166)-seeds and equivalent with HDR Ir(192)-brachytherapy were elaborated in MCNP5, shaped in a computational multivoxel simulator which reproduced a female thorax phantom. Spatial dose rate distributions and dose-volume histograms were generated. Protocol's analysis involving exposure time, seed's activities and dose were performed. Permanent Ho(166)-seed implants presented a maximum dose rate per unit of contained activity (MDR) of 1.1601 μGy h(-1) Bq(-1); and, a normalized MDR in standard points (8 mm, equidistant to 03-seeds - SP1, 10 mm - SP2) of 1.0% (SP1) and 0.5% (SP2), respectively. Ir(192)-brachytherapy presented MDR of 4.3945 × 10(-3) μGy h(-1) Bq(-1); and, 30% (SP1), and 20% (SP2). Therefore, seed's implant activities of 333 MBq (Ho(166)) and 259 GBq (Ir(192)) produced prescribed doses of 58 Gy (SP1; 5d) and 56 Gy (SP1, 5 fractions, 6 min), respectively. Breast Ho(166)-implants of 37-111 MBq are attractive due to the high dose rate near 6-10 mm from seeds, equivalent to Ir(192)-brachytherapy of 259 GBq (3 fractions, 6 min) providing similar dose in standard points at a week; however, with spatial dose distribution better confined. The seed positioning can be adjusted for controlling the breast tumor, in stages I and II, in flat and deep tumors, without any breast volumetric limitation.

  7. PSA bounce phenomenon after transperineal interstitial permanent prostate brachytherapy for localized prostate cancer

    International Nuclear Information System (INIS)

    Morita, Masashi; Lederer, J.L.; Fukagai, Takashi; Yoshida, Hideki; Shimada, Makoto

    2004-01-01

    We described the temporarily increase phenomenon in prostate-specific antigen level (PSA bounce) after transperineal interstitial permanent prostate brachytherapy (TIPPB) for localized prostate cancer. From December 1998 to May 2003, 500 consecutive patients with localized prostate cancer were treated with TIPPB using iodine-125 or palladium-103. We examined 200 patients who have more than 2-year PSA follow-up. Median follow-up length was 1,069 days (range, 712-1,411 days). No patient received neoadjuvant or adjuvant hormone therapy. PSA determinations were performed every 3 months for the first 2 years after procedure, and every 6 months hereafter. PSA bounce was defined as an increase of 0.1 ng/ml or greater above the preceding PSA level after implant followed by a subsequent decrease below that level. The American Society for Therapeutic Radiology and Oncology (ASTRO) consensus panel criteria 1996 were used to define biochemical failure. PSA bounce was observed in 40% (80/200) of the cases receiving TIPPB. The median time to PSA bounce was 13 months from the day of implant. The median magnitude of the PSA bounce was 0.3 ng/ml from the pre-bounce level. Twelve cases demonstrated biochemical failure according to the ASTRO consensus guidelines of three consecutive rises in PSA. Ten of these subsequently showed a drop in PSA, consistent with biologic control of their disease. Two cases remain classified as apparent biochemical failures. A transient rise in the PSA following TIPPB, the so-called ''bounce'' is a common occurrence. The apparent PSA control of ten of twelve cases failing by the ASTRO criteria raises some concern. Further observation will be necessary to determine ways to discriminate these from true disease progression. (author)

  8. Design and optimization of a brachytherapy robot

    Science.gov (United States)

    Meltsner, Michael A.

    Trans-rectal ultrasound guided (TRUS) low dose rate (LDR) interstitial brachytherapy has become a popular procedure for the treatment of prostate cancer, the most common type of non-skin cancer among men. The current TRUS technique of LDR implantation may result in less than ideal coverage of the tumor with increased risk of negative response such as rectal toxicity and urinary retention. This technique is limited by the skill of the physician performing the implant, the accuracy of needle localization, and the inherent weaknesses of the procedure itself. The treatment may require 100 or more sources and 25 needles, compounding the inaccuracy of the needle localization procedure. A robot designed for prostate brachytherapy may increase the accuracy of needle placement while minimizing the effect of physician technique in the TRUS procedure. Furthermore, a robot may improve associated toxicities by utilizing angled insertions and freeing implantations from constraints applied by the 0.5 cm-spaced template used in the TRUS method. Within our group, Lin et al. have designed a new type of LDR source. The "directional" source is a seed designed to be partially shielded. Thus, a directional, or anisotropic, source does not emit radiation in all directions. The source can be oriented to irradiate cancerous tissues while sparing normal ones. This type of source necessitates a new, highly accurate method for localization in 6 degrees of freedom. A robot is the best way to accomplish this task accurately. The following presentation of work describes the invention and optimization of a new prostate brachytherapy robot that fulfills these goals. Furthermore, some research has been dedicated to the use of the robot to perform needle insertion tasks (brachytherapy, biopsy, RF ablation, etc.) in nearly any other soft tissue in the body. This can be accomplished with the robot combined with automatic, magnetic tracking.

  9. Interstitially implanted I125 for prostate cancer using transrectal ultrasound

    International Nuclear Information System (INIS)

    Greenburg, S.; Petersen, J.; Hansen-Peters, I.; Baylinson, W.

    1990-01-01

    Prostate cancer is the third leading cause of death from cancer among men in the United States. Traditional treatments for prostate cancer are prostatectomy, external beam irradiation, and interstitial implantation of Iodine125 (I125) via laparotomy. These treatments are associated with significant morbidity and limitations. Based on experience with I125 interstitial implantation by transrectal ultrasound guidance for early-stage prostate cancer, it seems that this newer method of treatment has greater accuracy of placement and distribution of the isotope and has had few reported complications. The need for a surgical incision has been eliminated. Hospitalization time also has been decreased, creating the need for ambulatory and inpatient nurses to understand the importance of their respective roles in providing coordinated quality care for these patients. Nurses in these departments must have knowledge of the procedure, radiation safety, and common side effects related to the implant

  10. Dosimetric Consequences of Interobserver Variability in Delineating the Organs at Risk in Gynecologic Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Damato, Antonio L.; Townamchai, Kanopkis; Albert, Michele; Bair, Ryan J.; Cormack, Robert A.; Jang, Joanne; Kovacs, Arpad; Lee, Larissa J.; Mak, Kimberley S.; Mirabeau-Beale, Kristina L.; Mouw, Kent W.; Phillips, John G.; Pretz, Jennifer L.; Russo, Andrea L.; Lewis, John H.; Viswanathan, Akila N.

    2014-01-01

    Purpose: To investigate the dosimetric variability associated with interobserver organ-at-risk delineation differences on computed tomography in patients undergoing gynecologic interstitial brachytherapy. Methods and Materials: The rectum, bladder, and sigmoid of 14 patients treated with gynecologic interstitial brachytherapy were retrospectively contoured by 13 physicians. Geometric variability was calculated using κ statistics, conformity index (CI gen ), and coefficient of variation (CV) of volumes contoured across physicians. Dosimetric variability of the single-fraction D 0.1cc and D 2cc was assessed through CV across physicians, and the standard deviation of the total EQD2 (equivalent dose in 2 Gy per fraction) brachytherapy dose (SD TOT ) was calculated. Results: The population mean ± 1 standard deviation of κ, CI gen , and volume CV were, respectively: 0.77 ± 0.06, 0.70 ± 0.08, and 20% ± 6% for bladder; 0.74 ± 06, 0.67 ± 0.08, and 20% ± 5% for rectum; and 0.33 ± 0.20, 0.26 ± 0.17, and 82% ± 42% for sigmoid. Dosimetric variability was as follows: for bladder, CV = 31% ± 19% (SD TOT = 72 ± 64 Gy) for D 0.1cc and CV = 16% ± 10% (SD TOT = 9 ± 6 Gy) for D 2cc ; for rectum, CV = 11% ± 5% (SD TOT = 16 ± 17 Gy) for D 0.1cc and CV = 7% ± 2% (SD TOT = 4 ± 3 Gy) for D 2cc ; for sigmoid, CV = 39% ± 28% (SD TOT = 12 ± 18 Gy) for D 0.1cc and CV = 34% ± 19% (SD TOT = 4 ± 4 Gy) for D 2cc. Conclusions: Delineation of bladder and rectum by 13 physicians demonstrated substantial geometric agreement and resulted in good dosimetric agreement for all dose-volume histogram parameters except bladder D 0.1cc. Small delineation differences in high-dose regions by the posterior bladder wall may explain these results. The delineation of sigmoid showed fair geometric agreement. The higher dosimetric variability for sigmoid compared with rectum and bladder did not correlate with higher variability in the total brachytherapy dose but rather may be due to the

  11. SU-G-JeP2-14: MRI-Based HDR Prostate Brachytherapy: A Phantom Study for Interstitial Catheter Reconstruction with 0.35T MRI Images

    International Nuclear Information System (INIS)

    Park, S; Kamrava, M; Yang, Y

    2016-01-01

    Purpose: To evaluate the accuracy of interstitial catheter reconstruction with 0.35T MRI images for MRI-based HDR prostate brachytherapy. Methods: Recently, a real-time MRI-guided radiotherapy system combining a 0.35T MRI system and three cobalt 60 heads (MRIdian System, ViewRay, Cleveland, OH, USA) was installed in our department. A TrueFISP sequence for MRI acquisition at lower field on Viewray was chosen due to its fast speed and high signal-to-noise efficiency. Interstitial FlexiGuide needles were implanted into a tissue equivalent ultrasound prostate phantom (CIRS, Norfolk, Virginia, USA). After an initial 15s pilot MRI to confirm the location of the phantom, planning MRI was acquired with a 172s TrueFISP sequence. The pulse sequence parameters included: flip angle = 60 degree, echo time (TE) =1.45 ms, repetition time (TR) = 3.37 ms, slice thickness = 1.5 mm, field of view (FOV) =500 × 450mm. For a reference image, a CT scan was followed. The CT and MR scans were then fused with the MIM Maestro (MIM software Inc., Cleveland, OH, USA) and sent to the Oncentra Brachy planning system (Elekta, Veenendaal, Netherlands). Automatic catheter reconstruction using CT and MR image intensities followed by manual reconstruction was used to digitize catheters. The accuracy of catheter reconstruction was evaluated from the catheter tip location. Results: The average difference between the catheter tip locations reconstructed from the CT and MR in the transverse, anteroposterior, and craniocaudal directions was −0.1 ± 0.1 mm (left), 0.2 ± 0.2 mm (anterior), and −2.3 ± 0.5 mm (cranio). The average distance in 3D was 2.3 mm ± 0.5 mm. Conclusion: This feasibility study proved that interstitial catheters can be reconstructed with 0.35T MRI images. For more accurate catheter reconstruction which can affect final dose distribution, a systematic shift should be applied to the MR based catheter reconstruction in HDR prostate brachytherapy.

  12. Long-Term Outcome for Clinically Localized Prostate Cancer Treated With Permanent Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Taira, Al V.; Merrick, Gregory S.; Butler, Wayne M.; Galbreath, Robert W.; Lief, Jonathan; Adamovich, Edward; Wallner, Kent E.

    2011-01-01

    Purpose: To present the largest series of prostate cancer brachytherapy patients treated with modern brachytherapy techniques and postimplant day 0 dosimetric evaluation. Methods and Materials: Between April 1995 and July 2006, 1,656 consecutive patients were treated with permanent interstitial brachytherapy. Risk group stratification was carried out according to the Mt. Sinai guidelines. Median follow-up was 7.0 years. The median day 0 minimum dose covering at least 90% of the target volume was 118.8% of the prescription dose. Cause of death was determined for each deceased patient. Multiple clinical, treatment, and dosimetric parameters were evaluated for impact on the evaluated survival parameters. Results: At 12 years, biochemical progression-free survival (bPFS), cause-specific survival (CSS), and overall survival (OS) for the entire cohort was 95.6%, 98.2%, and 72.6%, respectively. For low-, intermediate-, and high-risk patients, bPFS was 98.6%, 96.5%, and 90.5%; CSS was 99.8%, 99.3%, and 95.2%; and OS was 77.5%, 71.1%, and 69.2%, respectively. For biochemically controlled patients, the median posttreatment prostate-specific antigen (PSA) concentration was 0.02 ng/ml. bPFS was most closely related to percent positive biopsy specimens and risk group, while Gleason score was the strongest predictor of CSS. OS was best predicted by patient age, hypertension, diabetes, and tobacco use. At 12 years, biochemical failure and cause-specific mortality were 1.8% and 0.2%, 5.1% and 2.1%, and 10.4% and 7.1% for Gleason scores 5 to 6 and 7 and ≥8, respectively. Conclusions: Excellent long-term outcomes are achievable with high-quality brachytherapy for low-, intermediate-, and high-risk patients. These results compare favorably to alternative treatment modalities including radical prostatectomy.

  13. Clinical outcome of high-dose-rate interstitial brachytherapy in patients with oral cavity cancer

    International Nuclear Information System (INIS)

    Lee, Sung Uk; Cho, Kwan Ho; Moon, Sung Ho; Choi, Sung Weon; Park, Joo Yong; Yun, Tak; Lee, Sang Hyun; Lim, Young Kyung; Jeong, Chi Young

    2014-01-01

    To evaluate the clinical outcome of high-dose-rate (HDR) interstitial brachytherapy (IBT) in patients with oral cavity cancer. Sixteen patients with oral cavity cancer treated with HDR remote-control afterloading brachytherapy using 192Ir between 2001 and 2013 were analyzed retrospectively. Brachytherapy was administered in 11 patients as the primary treatment and in five patients as salvage treatment for recurrence after the initial surgery. In 12 patients, external beam radiotherapy (50-55 Gy/25 fractions) was combined with IBT of 21 Gy/7 fractions. In addition, IBT was administered as the sole treatment in three patients with a total dose of 50 Gy/10 fractions and as postoperative adjuvant treatment in one patient with a total of 35 Gy/7 fractions. The 5-year overall survival of the entire group was 70%. The actuarial local control rate after 3 years was 84%. All five recurrent cases after initial surgery were successfully salvaged using IBT +/- external beam radiotherapy. Two patients developed local recurrence at 3 and 5 months, respectively, after IBT. The acute complications were acceptable (< or =grade 2). Three patients developed major late complications, such as radio-osteonecrosis, in which one patient was treated by conservative therapy and two required surgical intervention. HDR IBT for oral cavity cancer was effective and acceptable in diverse clinical settings, such as in the cases of primary or salvage treatment.

  14. Radiation protection procedures and dose to the staff in brachytherapy with permanent implant of the sources

    International Nuclear Information System (INIS)

    Tosi, G.; Cattani, F.

    2002-01-01

    The treatment of intra capsular prostate cancers with the permanent implantation of low energy sealed radioactive sources (''103 Pd-''125I) offers the same probability of curing the tumours as surgery and external-beam radiotherapy with a minimum incidence of unwanted side-effects. The first attempts of using sealed sources for treating prostate cancers go back to 1917, when Barringer reported the results obtained with the implant of ''236Ra needles. Beginning from that period the interest for prostate brachytherapy has shown a fluctuating trend, due especially to the technological possibilities and to the status of the alternative treatment modalities (surgery, external radiotherapy). The main reason of the substantial failure of brachytherapy as compared to the two other treatment modalities had two main causes: the energy, too high ( E≅ 840 keV), of γ-radiation emitted by ''226 Ra in equilibrium with its decay products and the lack of imaging techniques able to visualize with sufficient accuracy both the prostate and the arrangement, inside it, of the radioactive sources. The employ of low energy γ-emitting radionuclides began in 1974, when Whitmore et al. working at the Sloan Kettering Memorial Cancer Hospital of New York suggested the use of ''125 I sealed sources for the realisation of interstitial permanent implants. Also this attempt, though reducing the side effects typical of the surgical intervention (incontinence, impotence), did non give the expected results in terms of local control of the disease and, as a consequence, of the survival's length. This partial failure was attributed to the fact that, in most cases the dose distribution inside the target volume was not homogeneous, due to the inadequacy of the available imaging techniques used for checking the real position of the sources, during their manual insertion in the tissues. In the last ten years,however, great progresses have been made in the US i maging techniques, in the manufacture of

  15. Tumor hypoxia - A confounding or exploitable factor in interstitial brachytherapy? Effects of tissue trauma in an experimental rat tumor model

    NARCIS (Netherlands)

    van den Berg, AP; van Geel, CAJF; van Hooije, CMC; van der Kleij, AJ; Visser, AG

    2000-01-01

    Purpose: To evaluate the potential effects of tumor hypoxia induced by afterloading catheter implantation on the effectiveness of brachytherapy in a rat tumor model. Methods and Materials: Afterloading catheters (4) Here implanted in subcutaneously growing R1M rhabdomyosarcoma in female Wag/Rij

  16. Five-year biochemical outcome following permanent interstitial brachytherapy for clinical T1-T3 prostate cancer

    International Nuclear Information System (INIS)

    Merrick, Gregory S.; Butler, Wayne M.; Galbreath, Robert W.; Lief, Jonathan H.

    2001-01-01

    Purpose: To evaluate 5-year biochemical disease-free outcome for men with clinical T1b-T3a NxM0 1977 American Joint Committee on Cancer (1997 AJCC) adenocarcinoma of the prostate gland who underwent transperineal ultrasound-guided permanent prostate brachytherapy. Methods and Materials: Four hundred twenty-five patients underwent transperineal ultrasound-guided prostate brachytherapy using either 103 Pd or 125 I, for clinical T1b-T3a NxM0 (1997 AJCC) adenocarcinoma of the prostate gland, from April 1995 to October 1999. No patient underwent pathologic lymph-node staging. One hundred ninety patients were implanted with either 103 Pd or 125 I monotherapy; 235 patients received moderate-dose external beam radiation therapy (EBRT), followed by a prostate brachytherapy boost; 163 patients received neoadjuvant hormonal manipulation, in conjunction with either 103 Pd or 125 I monotherapy (77 patients) or in conjunction with moderate-dose EBRT and a prostate brachytherapy boost (86 patients). The median patient age was 68.0 years (range, 48.2-81.3 years). The median follow-up was 31 months (range, 11-69 months). Follow-up was calculated from the day of implantation. No patient was lost to follow-up. Biochemical disease-free survival was defined by the American Society of Therapeutic Radiation and Oncology (ASTRO) consensus definition. Results: For the entire cohort, the 5-year actuarial biochemical no evidence of disease (bNED) survival rate was 94%. For patients with low-, intermediate-, and high-risk disease, the 5-year biochemical disease-free rates were 97.1%, 97.5%, and 84.4%, respectively. For hormone-naive patients, 95.7%, 96.4%, and 79.9% of patients with low-, intermediate-, and high-risk disease were free of biochemical failure. Clinical and treatment parameters predictive of biochemical outcome included: clinical stage, pretreatment prostate-specific antigen (PSA), Gleason score, risk group, age > 65 years, and neoadjuvant hormonal therapy. Isotope choice was

  17. Toward a 3D transrectal ultrasound system for verification of needle placement during high-dose-rate interstitial gynecologic brachytherapy.

    Science.gov (United States)

    Rodgers, Jessica Robin; Surry, Kathleen; Leung, Eric; D'Souza, David; Fenster, Aaron

    2017-05-01

    Treatment for gynecologic cancers, such as cervical, recurrent endometrial, and vaginal malignancies, commonly includes external-beam radiation and brachytherapy. In high-dose-rate (HDR) interstitial gynecologic brachytherapy, radiation treatment is delivered via hollow needles that are typically inserted through a template on the perineum with a cylinder placed in the vagina for stability. Despite the need for precise needle placement to minimize complications and provide optimal treatment, there is no standard intra-operative image-guidance for this procedure. While some image-guidance techniques have been proposed, including magnetic resonance (MR) imaging, X-ray computed tomography (CT), and two-dimensional (2D) transrectal ultrasound (TRUS), these techniques have not been widely adopted. In order to provide intra-operative needle visualization and localization during interstitial brachytherapy, we have developed a three-dimensional (3D) TRUS system. This study describes the 3D TRUS system and reports on the system validation and results from a proof-of-concept patient study. To obtain a 3D TRUS image, the system rotates a conventional 2D endocavity transducer through 170 degrees in 12 s, reconstructing the 2D frames into a 3D image in real-time. The geometry of the reconstruction was validated using two geometric phantoms to ensure the accuracy of the linear measurements in each of the image coordinate directions and the volumetric accuracy of the system. An agar phantom including vaginal and rectal canals, as well as a model uterus and tumor, was designed and used to test the visualization and localization of the interstitial needles under idealized conditions by comparing the needles' positions between the 3D TRUS scan and a registered MR image. Five patients undergoing HDR interstitial gynecologic brachytherapy were imaged using the 3D TRUS system following the insertion of all needles. This image was manually, rigidly registered to the clinical

  18. SU-F-19A-12: Split-Ring Applicator with Interstitial Needle for Improved Volumetric Coverage in HDR Brachytherapy for Cervical Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Sherertz, T; Ellis, R; Colussi, V; Mislmani, M; Traughber, B; Herrmann, K; Podder, T [University Hospitals Case Medical Center, Cleveland, OH (United States)

    2014-06-15

    Purpose: To evaluate volumetric coverage of a Mick Radionuclear titanium Split-Ring applicator (SRA) with/without interstitial needle compared to an intracavitary Vienna applicator (VA), interstitial-intracavitary VA, and intracavitary ring and tandem applicator (RTA). Methods: A 57 year-old female with FIGO stage IIB cervical carcinoma was treated following chemoradiotherapy (45Gy pelvic and 5.4Gy parametrial boost) with highdose- rate (HDR) brachytherapy to 30Gy in 5 fractions using a SRA. A single interstitial needle was placed using the Ellis Interstitial Cap for the final three fractions to increase coverage of left-sided gross residual disease identified on 3T-MRI. High-risk (HR) clinical target volume (CTV) and intermediate-risk (IR) CTV were defined using axial T2-weighted 2D and 3D MRI sequences (Philips PET/MRI unit). Organs-at-risks (OARs) were delineated on CT. Oncentra planning system was used for treatment optimization satisfying GEC-ESTRO guidelines for target coverage and OAR constraints. Retrospectively, treatment plans (additional 20 plans) were simulated using intracavitary SRA (without needle), intracavitary VA (without needle), interstitial-intracavitary VA, and intracavitary RTA with this same patient case. Plans were optimized for each fraction to maintain coverage to HR-CTV. Results: Interstitial-intracavitary SRA achieved the following combined coverage for external radiation and brachytherapy (EQD2): D90 HR-CTV =94.6Gy; Bladder-2cc =88.9Gy; Rectum-2cc =65.1Gy; Sigmoid-2cc =48.9Gy; Left vaginal wall (VW) =103Gy, Right VW =99.2Gy. Interstitial-intracavitary VA was able to achieve identical D90 HR-CTV =94.6Gy, yet Bladder-2cc =91.9Gy (exceeding GEC-ESTRO recommendations of 2cc<90Gy) and Left VW =120.8Gy and Right VW =115.5Gy. Neither the SRA nor VA without interstitial needle could cover HR-CTV adequately without exceeding dose to Bladder-2cc. Conventional RTA was unable to achieve target coverage for the HR-CTV >80Gy without severely

  19. The role of interstitial brachytherapy in the treatment of vaginal and vulvar malignancies

    International Nuclear Information System (INIS)

    Seeger, A.R.; Windschall, A.; Lotter, M.; Sauer, R.; Strnad, V.; Mehlhorn, G.; Beckmann, M.W.

    2006-01-01

    Background: irradiation has established itself as a treatment for vulvar and vaginal malignancies. Due to the sensitive nature of the vulvar and vaginal tissues, interstitial brachytherapy (iBT) provides an effective, gentle and individualized therapy. Patients and Methods: patients with vulvar (nine of 22) and vaginal (13 of 22) malignancies were treated using interstitial pulsed-dose-rate brachytherapy (PDR-iBT). Twelve out of 22 patients were additionally treated using external-beam therapy to the pelvis and regional lymph nodes. The median total dose of PDR-iBT administered to patients with vulvar carcinoma was 55.0 Gy. The median total PDR dose administered to patients with vaginal malignancies amounted to 20.25 Gy. Results: the median follow-up time for patients with vulvar cancer was 19 months and for patients with vaginal malignancies 27 months. Acute mucositis or skin reactions during iBT were observed in 15 of 22 patients. Two of 22 patients showed delayed side effects. After 6 months, 77.8% of the patients with vulvar cancer (seven out of nine) and 100% of the patients with vaginal malignancies (13 out of 13) achieved complete local remission. One patient out of nine with vulvar carcinoma developed local recurrence, four out of nine regional recurrence, and two out of nine developed regional recurrence and had local tumor following therapy. In patients with malignancies of the vagina, no cases of local recurrence were observed, but distant metastases were found in five out of 13 patients. At the time of analysis, eleven out of 22 patients with vulvar or vaginal carcinoma were still alive. Conclusion: IBT achieved good local control without serious delayed side effects in both localizations. However, survival is limited by regional or distant metastases. (orig.)

  20. Dosimetric Consequences of Interobserver Variability in Delineating the Organs at Risk in Gynecologic Interstitial Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Damato, Antonio L., E-mail: adamato@lroc.harvard.edu [Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women' s Hospital, Boston, Massachusetts (United States); Townamchai, Kanopkis [Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women' s Hospital, Boston, Massachusetts (United States); Albert, Michele [Department of Radiation Oncology, Saint Anne' s Hospital Regional Cancer Center, Fall River, Massachusetts (United States); Bair, Ryan J. [Department of Radiology, Brigham and Women' s Hospital, Boston, Massachusetts (United States); Cormack, Robert A. [Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women' s Hospital, Boston, Massachusetts (United States); Jang, Joanne [Department of Radiation Oncology, Beth Israel Deaconess Medical Center, Boston, Massachusetts (United States); Kovacs, Arpad [Department of Radiology, Brigham and Women' s Hospital, Boston, Massachusetts (United States); Lee, Larissa J. [Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women' s Hospital, Boston, Massachusetts (United States); Mak, Kimberley S.; Mirabeau-Beale, Kristina L.; Mouw, Kent W.; Phillips, John G.; Pretz, Jennifer L.; Russo, Andrea L. [Harvard Radiation Oncology Program, Harvard Medical School, Boston, Massachusetts (United States); Lewis, John H.; Viswanathan, Akila N. [Department of Radiation Oncology, Dana-Farber Cancer Institute/Brigham and Women' s Hospital, Boston, Massachusetts (United States)

    2014-07-01

    Purpose: To investigate the dosimetric variability associated with interobserver organ-at-risk delineation differences on computed tomography in patients undergoing gynecologic interstitial brachytherapy. Methods and Materials: The rectum, bladder, and sigmoid of 14 patients treated with gynecologic interstitial brachytherapy were retrospectively contoured by 13 physicians. Geometric variability was calculated using κ statistics, conformity index (CI{sub gen}), and coefficient of variation (CV) of volumes contoured across physicians. Dosimetric variability of the single-fraction D{sub 0.1cc} and D{sub 2cc} was assessed through CV across physicians, and the standard deviation of the total EQD2 (equivalent dose in 2 Gy per fraction) brachytherapy dose (SD{sup TOT}) was calculated. Results: The population mean ± 1 standard deviation of κ, CI{sub gen}, and volume CV were, respectively: 0.77 ± 0.06, 0.70 ± 0.08, and 20% ± 6% for bladder; 0.74 ± 06, 0.67 ± 0.08, and 20% ± 5% for rectum; and 0.33 ± 0.20, 0.26 ± 0.17, and 82% ± 42% for sigmoid. Dosimetric variability was as follows: for bladder, CV = 31% ± 19% (SD{sup TOT} = 72 ± 64 Gy) for D{sub 0.1cc} and CV = 16% ± 10% (SD{sup TOT} = 9 ± 6 Gy) for D{sub 2cc}; for rectum, CV = 11% ± 5% (SD{sup TOT} = 16 ± 17 Gy) for D{sub 0.1cc} and CV = 7% ± 2% (SD{sup TOT} = 4 ± 3 Gy) for D{sub 2cc}; for sigmoid, CV = 39% ± 28% (SD{sup TOT} = 12 ± 18 Gy) for D{sub 0.1cc} and CV = 34% ± 19% (SD{sup TOT} = 4 ± 4 Gy) for D{sub 2cc.} Conclusions: Delineation of bladder and rectum by 13 physicians demonstrated substantial geometric agreement and resulted in good dosimetric agreement for all dose-volume histogram parameters except bladder D{sub 0.1cc.} Small delineation differences in high-dose regions by the posterior bladder wall may explain these results. The delineation of sigmoid showed fair geometric agreement. The higher dosimetric variability for sigmoid compared with rectum and bladder did not correlate with

  1. The study of the radiation protection problem in the radionuclide interstitial implantation therapy

    International Nuclear Information System (INIS)

    Zhang Jimian

    2006-01-01

    Objective: To analyze and study the radiation protection problem in the radionuclide interstitial permanent implantation therapy. Methods: Based on test data from radioactive measurement department, calculating results and national standards, the radiation dose of the exposed radioactive particles, the operator who has participated in the radionuclide interstitial permanent implantation therapy operation and the relatives who have accompanied the patient during the whole course, the reference time of being discharged from hospital for the patients who have been cured by different activity of radioactive particles are studied. Results: The maximal radiation dose of operating doctor who has participated in a single radionuclide interstitial permanent implantation therapy operation and the relatives who has accompanied the patient during the whole course are 0.315 mSv/a and 0.70 mSv. Based on actual contact frequencies, their radiation dose is proved to be smaller than the restricted dose prescribed by national standards. The reference time of leaving hospital for the patients who have been cured by different activity of radioactive particles is 0 to 44 days. Conclusion: The radiation dose of radiation workers and surrounding publics in the radionuclide interstitial permanent implantation therapy operation can be acceptable under certain shields. But the risk of potential exposure should be guarded. The authors should Lay down operation indications and avoid performing operation blindly. If one must be operated, the authors should plan the quantity and the part of the painting radioactive particles accurately in order to avoid some passible complications. (authors)

  2. Implants with 32P-foils for LDR-brachytherapy of benign stenosis in urology and gastroenterology

    International Nuclear Information System (INIS)

    Assmann, Walter; Becker, Ricarda; Otto, Henrike

    2013-01-01

    For LDR-brachytherapy, a limited number of implant geometries and materials are available. To avoid wound healing related hyper-proliferation (stenosis, keloids) a novel radioactive foil system was developed based on beta emitting 32 P, which can be easily integrated in existing implants such as urethral catheters or bile duct stents. As substrate material for these foils PEEK (polyetherethercetone) was chosen because of its radiation hardness during neutron activation of 32 P. The activity was determined by liquid scintillation counting and gamma spectroscopy, dose distributions were measured with scintillation detectors and radiochromic films. The correlation between activity and dose was checked by Monte-Carlo-simulations (Geant4). Prototypes of the 32 P-implants have shown in wash-out tests the required tightness for sealed radioactive sources. In animal tests on urethra and bile duct, the uncomplicated and save application of 32 P-foils mounted on standard implants has been demonstrated, which is almost unchanged due to the simple radiation protection with plexiglass. This concept of radioactive implants with integrated 32 P-foils could extend essentially the application possibilities of LDR-brachytherapy. (orig.)

  3. Dosimetric comparison of different dose prescription systems with CT based intracavitary brachytherapy and manual back projection technique to reconstruct the applicator

    International Nuclear Information System (INIS)

    Oinam, A.S.; Dubey, S.; Kehwar, T.S.; Rout, Sanjaya K.; Patel, F.D.; Sharma, S.C.; Goyal, D.R.; Narayan, P.

    2002-01-01

    Intracavitary brachytherapy is one of the well-established techniques for the treatment of carcinoma of cervix. The prediction of late effect of normal tissue like rectum and bladder needs the defining of the volume of the bladder and rectum in situ. In the normal planning of intracavitary and interstitial implants, simulated radiograph films are used to reconstruct the applicator geometry and dose points to represent the dose to critical organs. CT based brachytherapy can define such volume instead of defining dose points, which represent the dose to these critical organs

  4. High dose rate brachytherapy in treatment of high grade astrocytomas

    International Nuclear Information System (INIS)

    Garcia-Alejo, R.; Delgado, J.M.; Cerro, E. del; Torres, J.J.; Martinez, R.

    1996-01-01

    From May 1994 to June 1995, 18 patients with high grade astrocytomas were entered prospectively on a selective protocol combining surgery, external beam radiotherapy, stereotactic interstitial implantation with HDR Iridium 192 and chemotherapy. Only those patients with tumor size 100cc or less average dimension, high grade astrocytoma, Karnofsky 70 or greater, unilateral, circumscribed, unifocal, tumor stable or responding to external radiation and supratentorial were included in the study. Ages ranged from 16 to 69 years. There were 13 males and 5 females. Surgery consisted of biopsy only in 3 patients, subtotal resection in 11, and gross total resection in 4 patients. Focal external beam radiation portals included the contrast enhancing mass on CT scan plus a 3 cm margin. The protocol called for minimum tumor dose of 60 Gy to be given in 2 Gy daily fractions. An interstitial brachytherapy boost was to be performed two weeks after the conclusion of external beam radiation. The dose was 30 Gy in 4 fractions. The authors analyze on basis on their personal experience, the possibilities and the limits offered by this therapeutic procedure in neuro-oncology. Using stereotactic techniques, interstitial brachytherapy of brain tumors was technically possible with negligible acute morbidity and mortality, and appeared to be effective and may provide for an increase in tumor control in selected cases

  5. Seed Implant Retention Score Predicts the Risk of Prolonged Urinary Retention After Prostate Brachytherapy

    International Nuclear Information System (INIS)

    Lee, Hoon K.; Adams, Marc T.; Shi, Qiuhu; Basillote, Jay; LaMonica, Joanne; Miranda, Luis; Motta, Joseph

    2010-01-01

    Purpose: To risk-stratify patients for urinary retention after prostate brachytherapy according to a novel seed implant retention score (SIRS). Patients and Methods: A total of 835 patients underwent transperineal prostate seed implant from March 1993 to January 2007; 197 patients had 125 I and 638 patients had 103 Pd brachytherapy. Four hundred ninety-four patients had supplemental external-beam radiation. The final downsized prostate volume was used for the 424 patients who had neoadjuvant hormone therapy. Retention was defined as reinsertion of a Foley catheter after the implant. Results: Retention developed in 7.4% of patients, with an average duration of 6.7 weeks. On univariate analysis, implant without supplemental external-beam radiation (10% vs. 5.6%; p = 0.02), neoadjuvant hormone therapy (9.4% vs. 5.4%; p = 0.02), baseline α-blocker use (12.5% vs. 6.3%; p = 0.008), and increased prostate volume (13.4% vs. 6.9% vs. 2.9%, >45 cm 3 , 25-45 cm 3 , 3 ; p = 0.0008) were significantly correlated with increased rates of retention. On multivariate analysis, implant without supplemental external-beam radiation, neoadjuvant hormone therapy, baseline α-blocker use, and increased prostate volume were correlated with retention. A novel SIRS was modeled as the combined score of these factors, ranging from 0 to 5. There was a significant correlation between the SIRS and retention (p < 0.0001). The rates of retention were 0, 4%, 5.6%, 9%, 20.9%, and 36.4% for SIRS of 0 to 5, respectively. Conclusions: The SIRS may identify patients who are at high risk for prolonged retention after prostate brachytherapy. A prospective validation study of the SIRS is planned.

  6. Radiological protection on interstitial brachytherapy and dose determination and exposure rate of an Ir-192 source through the MCNP-4B; Proteccion radiologica en braquiterapia intersticial y determinacion de la dosis y tasa de exposicion de una fuente de Ir-192 mediante el MCNP-4B

    Energy Technology Data Exchange (ETDEWEB)

    Morales L, M.E. [INEN, Av. Angamos Este 2520- Surquillo, Lima (Peru)

    2006-07-01

    The present work was carried out in the Neurological Sciences Institute having as objective to determine the dose and the rate of exhibition of the sources of Iridium 192, Iodine 125 and Palladium 103; which are used to carry out implant in the Interstitial Brachytherapy according to the TG43. For it we carry out a theoretical calculation, its are defined in the enter file: the geometry, materials of the problem and the radiation source, etc; in the MCNP-4B Monte Carlo code, considering a punctual source and for the dose determination we simulate thermoluminescent dosemeters (TLD): at 5 cm, 50 cm, 100 cm and 200 cm of the source. Our purpose is to analyze the radioprotection measures that should take into account in this Institute in which are carried out brain biopsies using a Micro mar stereotactic mark, and in a near future with the collaboration of a doctor and a cuban physique seeks to be carried out the Interstitial Brachytherapy technique with sources of Ir-192 for patient with tumors like glioblastoma, astrocytoma, etc. (Author)

  7. Implants with {sup 32}P-foils for LDR-brachytherapy of benign stenosis in urology and gastroenterology; {sup 32}P-haltige Folien als Implantate fuer die LDR-Brachytherapie gutartiger Stenosen in der Urologie und Gastroenterologie

    Energy Technology Data Exchange (ETDEWEB)

    Assmann, Walter [Ludwig-Maximilians-Univ., Muenchen (Germany). Fakultaet fuer Physik; Becker, Ricarda; Otto, Henrike [Klinikum der Universitaet Muenchen, Campus Grosshadern (Germany). Laser-Forschungslabor] [and others

    2013-03-01

    For LDR-brachytherapy, a limited number of implant geometries and materials are available. To avoid wound healing related hyper-proliferation (stenosis, keloids) a novel radioactive foil system was developed based on beta emitting {sup 32}P, which can be easily integrated in existing implants such as urethral catheters or bile duct stents. As substrate material for these foils PEEK (polyetherethercetone) was chosen because of its radiation hardness during neutron activation of {sup 32}P. The activity was determined by liquid scintillation counting and gamma spectroscopy, dose distributions were measured with scintillation detectors and radiochromic films. The correlation between activity and dose was checked by Monte-Carlo-simulations (Geant4). Prototypes of the {sup 32}P-implants have shown in wash-out tests the required tightness for sealed radioactive sources. In animal tests on urethra and bile duct, the uncomplicated and save application of {sup 32}P-foils mounted on standard implants has been demonstrated, which is almost unchanged due to the simple radiation protection with plexiglass. This concept of radioactive implants with integrated {sup 32}P-foils could extend essentially the application possibilities of LDR-brachytherapy. (orig.)

  8. Patient-based dosimetric comparison of interstitial and intracavitary brachytherapy in cases of cancer cervix

    International Nuclear Information System (INIS)

    Bansal, Anil K.; Julka, P.K.; Sharma, D.N.; Rustogi, Ashish; Subramani, V.; Prabhakar, R.; Rath, G.K.; Semwal, Manoj K.; Thulkar, S.

    2008-01-01

    Brachytherapy in the form of High Dose Rate (HDR) intracavitary radiotherapy (ICRT) along with external beam radiotherapy(EBRT) is the main treatment in cancer cervix. Of late, some large centres have started practicing template based transperineal interstitial brachytherapy (TIB) for advanced/ bulky cancer cervix. Usually, TIB is given for patients with advanced disease/ distorted anatomy or recurrent disease for better lateral target coverage. CT/MRI/USG based planning has made volumetric dosimetry possible for the target and the organs at risk (OARs). This has resulted in better correlation between dose received and treatment outcome in terms of tumour control and late toxicities as against the point dosimetry system. It has been shown by many studies that ICRU based point dose reporting may not represent the actual doses received by the OARs. Though it is expected that TIB gives better target coverage and OAR sparing in advanced/ bulky cancer cervix cases as compared to ICRT, detailed patient studies on the subject have not been reported. We have carried out dosimetric comparison between ICRT and TIB for cancer cervix patients undergoing treatment at our centre in terms of treated volume and doses to OARs

  9. Retrospective study of 87 interstitial brachytherapy for facial skin carcinomas

    International Nuclear Information System (INIS)

    Berland, E.; Bolla, M.; Saillard, F.; Beani, J.C.; Lebeau, J.; Fillon, J.P.; Vrousos, C.

    1996-01-01

    From 1977 to 1983, 83 patients (pts) with 87 facial skin carcinomas (FSC) were treated with interstitial brachytherapy (BT) in our department. Median age was 70 years (43-97 y). There was 61 basal cell carcinomas and 26 squamous cell carcinomas. 28 of them corresponded to local relapse of FSC initially treated without BT. TNM stadification was as following: T1+rT1=72,4%; T2+rT2=23%; T3+rT3=2,3%; T4+rT4=2,3%. The median follow-up was 53 months (8 months to 15 years). Different techniques were used:hypodermic needles,plastic tubes, Raynal method. A dose of 60 Gy was prescribed on the reference isodose, according to the PARIS system. Specific overall survival was 98,8% at 5 and 10 y. Local control at 5 and 10 y was 100 and 90% for FSC treated in first intention, and 81,2% for recurrent FSC. This was highly significant (p=0,002 Logrank test) Functional sequelae were appreciated according to the RTOG/EORTC scale: 29 occurred in 21 pts.60% of them were depigmentation ((11(29))) and epilation ((6(29))). Among the 21 FSC of the eyelid, one lacrimal stenosis occurred (4,8%) Cosmetic results were judged 'good' or 'very good' for 83% of the pts (BONVALLOT criteria). Bad results occured more frequently for recurrent FSC According to the litterature, our study shows that interstitial BT is a good alternative to surgery for facial skin carcinoma for elderly patients

  10. 78 FR 41125 - Interim Enforcement Policy for Permanent Implant Brachytherapy Medical Event Reporting

    Science.gov (United States)

    2013-07-09

    ... NUCLEAR REGULATORY COMMISSION [NRC-2013-0114] Interim Enforcement Policy for Permanent Implant Brachytherapy Medical Event Reporting AGENCY: Nuclear Regulatory Commission. ACTION: Policy statement; revision. SUMMARY: The U.S. Nuclear Regulatory Commission (NRC) is issuing an interim Enforcement Policy that allows...

  11. The effects of metallic implants on electroporation therapies: feasibility of irreversible electroporation for brachytherapy salvage.

    Science.gov (United States)

    Neal, Robert E; Smith, Ryan L; Kavnoudias, Helen; Rosenfeldt, Franklin; Ou, Ruchong; Mclean, Catriona A; Davalos, Rafael V; Thomson, Kenneth R

    2013-12-01

    Electroporation-based therapies deliver brief electric pulses into a targeted volume to destabilize cellular membranes. Nonthermal irreversible electroporation (IRE) provides focal ablation with effects dependent on the electric field distribution, which changes in heterogeneous environments. It should be determined if highly conductive metallic implants in targeted regions, such as radiotherapy brachytherapy seeds in prostate tissue, will alter treatment outcomes. Theoretical and experimental models determine the impact of prostate brachytherapy seeds on IRE treatments. This study delivered IRE pulses in nonanimal, as well as in ex vivo and in vivo tissue, with and in the absence of expired radiotherapy seeds. Electrical current was measured and lesion dimensions were examined macroscopically and with magnetic resonance imaging. Finite-element treatment simulations predicted the effects of brachytherapy seeds in the targeted region on electrical current, electric field, and temperature distributions. There was no significant difference in electrical behavior in tissue containing a grid of expired radiotherapy seeds relative to those without seeds for nonanimal, ex vivo, and in vivo experiments (all p > 0.1). Numerical simulations predict no significant alteration of electric field or thermal effects (all p > 0.1). Histology showed cellular necrosis in the region near the electrodes and seeds within the ablation region; however, there were no seeds beyond the ablation margins. This study suggests that electroporation therapies can be implemented in regions containing small metallic implants without significant changes to electrical and thermal effects relative to use in tissue without the implants. This supports the ability to use IRE as a salvage therapy option for brachytherapy.

  12. Clinical feasibility of combined intracavitary/interstitial brachytherapy in locally advanced cervical cancer employing MRI with a tandem/ring applicator in situ and virtual preplanning of the interstitial component

    International Nuclear Information System (INIS)

    Fokdal, Lars; Tanderup, Kari; Hokland, Steffen Bjerre; Røhl, Lisbeth; Pedersen, Erik Morre; Nielsen, Søren Kynde; Paludan, Merete; Lindegaard, Jacob Christian

    2013-01-01

    Purpose: To investigate the reproducibility of virtually planned needles, changes in DVH parameters and clinical feasibility of combined intracavitary/interstitial (IC/IS) pulsed dose rate brachytherapy (PDR-BT) for locally advanced cervical cancer based on 3D MRI preplanning. Material and methods: Fifty-eight consecutively patients accrued in the EMBRACE study were included. Treatment was initiated with external beam radiotherapy and cisplatin. Three BT implants and MRI with the applicator in situ were performed in all patients, i.e. week 5 (BT0), week 6 (BT1) and week 7 (BT2) of the treatment. BT0 was only used for preplanning of subsequent implantations, whereas BT1 and BT2 comprised 2 equal sized fractions of PDR BT. Results: Based on BT0, 24 patients (41%) were selected for a combined IC/IS implant at BT1 and BT2. Patients treated with IC/IS BT had significantly larger tumours compared with patients treated with IC BT only (p < 0.03). Additional time in general anaesthesia for the IC/IS component was on average 16 min. The number of preplanned virtual needles was 5.3 ± 2.7 compared to 5.3 ± 2.9 and 5.4 ± 3.0 needles implanted at BT1 and BT2, respectively (p = 0.72). Planned needle implantation depth was 33 ± 15 mm compared to 30 ± 10 mm at BT1 and 29 ± 11 mm at BT2 (p = 0.04). In the 24 patients selected for IC/IS BT both the virtual IC/IS plan (BT0) and the actually delivered plan (BT1 + BT2) significantly increased D90 and D100 for HR CTV (p < 0.01) and reduced D2cc for sigmoid (p < 0.01) and bowel (p = 0.04) compared to the optimised IC preplan (BT0). IC/IS BT was only associated with minor morbidity, which was resolved at a 3-month follow up. Conclusion: Combined IC/IS BT based on full 3D MRI preplanning is clinically feasible. The virtual preplanned needle positions are reproducible at subsequent BT applications leading to significantly improved DVH parameters and a clinically feasible and fast implant procedure

  13. Successful treatment of a 67-year-old woman with urethral adenocarcinoma with the use of external beam radiotherapy and image guided adaptive interstitial brachytherapy

    DEFF Research Database (Denmark)

    Mujkanovic, Jasmin; Tanderup, Kari; Agerbæk, Mads

    2016-01-01

    Primary urethral cancer (PUC) is a very rare disease. This case report illustrates a successful treatment approach of a 67-year-old woman with a urethral adenocarcinoma selected for an organ preserving treatment with external beam radiotherapy (EBRT) and interstitial brachytherapy (BT) boost, using...

  14. AAPM and GEC-ESTRO guidelines for image-guided robotic brachytherapy: Report of Task Group 192

    Energy Technology Data Exchange (ETDEWEB)

    Podder, Tarun K., E-mail: tarun.podder@uhhospitals.org [Department of Radiation Oncology, University Hospitals, Case Western Reserve University, Cleveland, Ohio 44122 (United States); Beaulieu, Luc [Department of Radiation Oncology, Centre Hospitalier Univ de Quebec, Quebec G1R 2J6 (Canada); Caldwell, Barrett [Schools of Industrial Engineering and Aeronautics and Astronautics, Purdue University, West Lafayette, Indiana 47907 (United States); Cormack, Robert A. [Department of Radiation Oncology, Harvard Medical School, Boston, Massachusetts 02115 (United States); Crass, Jostin B. [Department of Radiation Oncology, Vanderbilt University, Nashville, Tennessee 37232 (United States); Dicker, Adam P.; Yu, Yan [Department of Radiation Oncology, Thomas Jefferson University, Philadelphia, Pennsylvania 19107 (United States); Fenster, Aaron [Department of Imaging Research, Robarts Research Institute, London, Ontario N6A 5K8 (Canada); Fichtinger, Gabor [School of Computer Science, Queen’s University, Kingston, Ontario K7L 3N6 (Canada); Meltsner, Michael A. [Philips Radiation Oncology Systems, Fitchburg, Wisconsin 53711 (United States); Moerland, Marinus A. [Department of Radiotherapy, University Medical Center Utrecht, Utrecht, 3508 GA (Netherlands); Nath, Ravinder [Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut 06520 (United States); Rivard, Mark J. [Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111 (United States); Salcudean, Tim [Department of Electrical and Computer Engineering, University of British Columbia, Vancouver, British Columbia V6T 1Z4 (Canada); Song, Danny Y. [Department of Radiation Oncology, Johns Hopkins University School of Medicine, Baltimore, Maryland 21231 (United States); Thomadsen, Bruce R. [Department of Medical Physics, University of Wisconsin, Madison, Wisconsin 53705 (United States)

    2014-10-15

    In the last decade, there have been significant developments into integration of robots and automation tools with brachytherapy delivery systems. These systems aim to improve the current paradigm by executing higher precision and accuracy in seed placement, improving calculation of optimal seed locations, minimizing surgical trauma, and reducing radiation exposure to medical staff. Most of the applications of this technology have been in the implantation of seeds in patients with early-stage prostate cancer. Nevertheless, the techniques apply to any clinical site where interstitial brachytherapy is appropriate. In consideration of the rapid developments in this area, the American Association of Physicists in Medicine (AAPM) commissioned Task Group 192 to review the state-of-the-art in the field of robotic interstitial brachytherapy. This is a joint Task Group with the Groupe Européen de Curiethérapie-European Society for Radiotherapy and Oncology (GEC-ESTRO). All developed and reported robotic brachytherapy systems were reviewed. Commissioning and quality assurance procedures for the safe and consistent use of these systems are also provided. Manual seed placement techniques with a rigid template have an estimated in vivo accuracy of 3–6 mm. In addition to the placement accuracy, factors such as tissue deformation, needle deviation, and edema may result in a delivered dose distribution that differs from the preimplant or intraoperative plan. However, real-time needle tracking and seed identification for dynamic updating of dosimetry may improve the quality of seed implantation. The AAPM and GEC-ESTRO recommend that robotic systems should demonstrate a spatial accuracy of seed placement ≤1.0 mm in a phantom. This recommendation is based on the current performance of existing robotic brachytherapy systems and propagation of uncertainties. During clinical commissioning, tests should be conducted to ensure that this level of accuracy is achieved. These tests

  15. AAPM and GEC-ESTRO guidelines for image-guided robotic brachytherapy: Report of Task Group 192

    International Nuclear Information System (INIS)

    Podder, Tarun K.; Beaulieu, Luc; Caldwell, Barrett; Cormack, Robert A.; Crass, Jostin B.; Dicker, Adam P.; Yu, Yan; Fenster, Aaron; Fichtinger, Gabor; Meltsner, Michael A.; Moerland, Marinus A.; Nath, Ravinder; Rivard, Mark J.; Salcudean, Tim; Song, Danny Y.; Thomadsen, Bruce R.

    2014-01-01

    In the last decade, there have been significant developments into integration of robots and automation tools with brachytherapy delivery systems. These systems aim to improve the current paradigm by executing higher precision and accuracy in seed placement, improving calculation of optimal seed locations, minimizing surgical trauma, and reducing radiation exposure to medical staff. Most of the applications of this technology have been in the implantation of seeds in patients with early-stage prostate cancer. Nevertheless, the techniques apply to any clinical site where interstitial brachytherapy is appropriate. In consideration of the rapid developments in this area, the American Association of Physicists in Medicine (AAPM) commissioned Task Group 192 to review the state-of-the-art in the field of robotic interstitial brachytherapy. This is a joint Task Group with the Groupe Européen de Curiethérapie-European Society for Radiotherapy and Oncology (GEC-ESTRO). All developed and reported robotic brachytherapy systems were reviewed. Commissioning and quality assurance procedures for the safe and consistent use of these systems are also provided. Manual seed placement techniques with a rigid template have an estimated in vivo accuracy of 3–6 mm. In addition to the placement accuracy, factors such as tissue deformation, needle deviation, and edema may result in a delivered dose distribution that differs from the preimplant or intraoperative plan. However, real-time needle tracking and seed identification for dynamic updating of dosimetry may improve the quality of seed implantation. The AAPM and GEC-ESTRO recommend that robotic systems should demonstrate a spatial accuracy of seed placement ≤1.0 mm in a phantom. This recommendation is based on the current performance of existing robotic brachytherapy systems and propagation of uncertainties. During clinical commissioning, tests should be conducted to ensure that this level of accuracy is achieved. These tests

  16. [Implants with 32P-foils for LDR-brachytherapy of benign stenosis in urology and gastroenterology].

    Science.gov (United States)

    Assmann, Walter; Becker, Ricarda; Otto, Henrike; Bader, Markus; Clemente, Lucas; Reinhardt, Sabine; Schäfer, Claus; Schirra, Jörg; Uschold, Stephanie; Welzmüller, Andreas; Sroka, Ronald

    2013-02-01

    For LDR-brachytherapy, a limited number of implant geometries and materials are available. To avoid wound healing related hyper-proliferation (stenosis, keloids) a novel radioactive foil system was developed based on beta emitting (32)P, which can be easily integrated in existing implants such as urethral catheters or bile duct stents. As substrate material for these foils PEEK (polyetherethercetone) was chosen because of its radiation hardness during neutron activation of (32)P. The activity was determined by liquid scintillation counting and gamma spectroscopy, dose distributions were measured with scintillation detectors and radiochromic films. The correlation between activity and dose was checked by Monte-Carlo-simulations (Geant4). Prototypes of the (32)P-implants have shown in wash-out tests the required tightness for sealed radioactive sources. In animal tests on urethra and bile duct, the uncomplicated and save application of (32)P-foils mounted on standard implants has been demonstrated, which is almost unchanged due to the simple radiation protection with plexiglass. This concept of radioactive implants with integrated (32)P-foils could extend essentially the application possibilities of LDR-brachytherapy. Copyright © 2012. Published by Elsevier GmbH.

  17. Randomized comparison between intracoronary β-radiation brachytherapy and implantation of paclitaxel-eluting stents for the treatment of diffuse in-stent restenosis

    International Nuclear Information System (INIS)

    Schukro, Christoph; Syeda, Bonni; Kirisits, Christian; Schmid, Rainer; Pichler, Philipp; Pokrajac, Boris; Lang, Irene; Poetter, Richard; Glogar, Dietmar

    2007-01-01

    Background and purpose: Intracoronary brachytherapy was the primary therapeutic option for the treatment of in-stent restenosis (ISR) during the last years. Especially for the treatment of diffuse ISR (lesions >10 mm), β-source brachytherapy was significantly superior to singular balloon angioplasty. Despite lacking clinical database, the implantation of drug eluting stents recently became a common procedure for the treatment of ISR. This randomized trial aimed to compare the efficacy of β-brachytherapy with β-radioisotopes 90 Sr/ 90 Y and paclitaxel-eluting stent implantation for the treatment of diffuse ISR. Material and methods: Thirty-seven patients with diffuse ISR were randomly assigned to β-brachytherapy after balloon angioplasty (Beta-Cath TM in 17 patients) or paclitaxel-eluting stent implantation (Taxus-Express2 TM in 20 patients). Six-month clinical follow-up was obtained for all patients, while angiographic follow-up was available for 30 patients. Results: Binary ISR (restenosis >50%) within target segment was observed in three patients treated with Beta-Cath TM , of which one needed target segment revascularisation for recurrent ISR, whereas no significant restenosis occurred in the patients treated with Taxus-Express2 TM (P = 0.037). No further major adverse cardiac (target segment revascularisation, myocardial infarction, death) was found in either group (P = NS). Stent implantation was the more time-saving (31 ± 11 min versus 60 ± 23 min, P TM arm, we found no difference in clinical outcome after implantation of paclitaxel-eluting stents for the treatment of diffuse ISR when compared to β-brachytherapy

  18. Dose and volume specification for reporting interstitial therapy

    International Nuclear Information System (INIS)

    1997-01-01

    The ICRU has previously published reports dealing with Dose Specification for Reporting External Beam Therapy with Photons and Electrons (ICRU Report 29, ICRU, 1978), Dose Specification for Reporting External Beam Therapy (ICRU Report 50, ICRU, 1993) and Dose and Volume Specification for Reporting Intracavitary Therapy in Gynecology (ICRU Report 38, ICRU, 1985). The present report addresses the problem of absorbed dose specification for report interstitial therapy. Although specific to interstitial therapy, many of the concepts developed in this report are also applicable to certain other kinds of brachytherapy applications. In particular, special cases of intraluminal brachytherapy and plesio-brachytherapy via surface molds employing x or gamma emitters are addressed in this report

  19. Suppression of self-interstitials in silicon during ion implantation via in-situ photoexcitation

    International Nuclear Information System (INIS)

    Ravi, J.; Erokhin, Yu.; Christensen, K.; Rozgonyi, G.A.; Patnaik, B.K.; White, C.W.

    1995-02-01

    The influence of in-situ photoexcitation during low temperature implantation on self-interstitial agglomeration following annealing has been investigated using transmission electron microscopy (TEM). A reduction in the level of as-implanted damage determined by RBS and TEM occurs athermally during 150 keV self-ion implantation. The damage reduction following a 300 C anneal suggests that it is mostly divacancy related. Subsequent thermal annealing at 800 C resulted in the formation of (311) rod like defects or dislocation loops for samples with and without in-situ photoexcitation, respectively. Estimation of the number of self-interstitials bound by these defects in the sample without in-situ photoexcitation corresponds to the implanted dose; whereas for the in-situ photoexcitation sample a suppression of ∼2 orders in magnitude is found. The kinetics of the athermal annealing process are discussed within the framework of either a recombination enhanced defect reaction mechanism, or a charge state enhanced defect migration and Coulomb interaction

  20. Iodine-125 orbital brachytherapy with a prosthetic implant in situ

    Energy Technology Data Exchange (ETDEWEB)

    Stannard, Clare [Groote Schuur Hospital and Cape Town Univ. (South Africa). Dept. of Radiation Oncology; Maree, Gert; Munro, Roger [Groote Schuur Hospital and Cape Town Univ. (South Africa). Dept. of Medical Physics; Lecuona, Karin [Groote Schuur Hospital and Cape Town Univ. (South Africa). Dept. of Ophthalmology; Sauerwein, Wolfgang [Universitaetsklinikum Essen (Germany). Strahlenklinik, NCTeam

    2011-05-15

    Purpose: Brachytherapy is one method of irradiating the orbit after enucleation of an eye with a malignant tumor that has a potential to recur. It consists of 6 trains of I-125 seeds placed around the periphery of the orbit, a shorter central train, and a metal disc, loaded with seeds, placed beneath the eyelids. The presence of a prosthetic orbital implant requires omission of the central train and adjustment of the activity of the seeds in the anterior orbit around the prosthesis. Patients and Methods: This is a retrospective review of the technical modifications and outcome of 12 patients treated in this manner: 6 with retinoblastoma, 5 with malignant melanoma, and 1 with an intraocular rhabdomyosarcoma. The median dose was 35.5 Gy in 73 hours for retinoblastoma and 56 Gy in 141 hours for malignant melanoma. Patients with retinoblastoma and rhabdomyosarcoma also received chemotherapy. Results: The tubes can be placed satisfactorily around the prosthesis. The increased activity in the anterior half of the tubes produced comparable dose distributions. There have been no orbital recurrences, no extrusion of the prosthesis, and cosmesis is good. Conclusion: Insertion of a prosthetic implant at the time of enucleation greatly enhances the subsequent cosmetic appearance. This should be encouraged unless there is frank tumor in the orbit. Orbital brachytherapy without the central train continues to give excellent local control. The short treatment time and good cosmesis are added advantages. The patient is spared the expense and inconvenience of removing and replacing the prosthetic implant. (orig.)

  1. Late change of normal tissue treated either by high dose rate or low dose rate interstitial brachytherapy. A retrospective comparative study on oral and oropharyngeal mucosa

    International Nuclear Information System (INIS)

    Nose, Takayuki; Koizumi, Masahiko; Nishiyama, Kinji; Inoue, Toshihiko

    2002-01-01

    The purpose of this study was to compare late changes of normal tissue treated either by high dose rate (HDR) or low dose rate (LDR) interstitial brachytherapy. For HDR group, 22 oropharynx cancer patients who were treated by HDR Ir-192 interstitial brachytherapy with/without external beam radiotherapy in Osaka (Osaka Medical Center for Cancer and Cardiovascular Diseases and Osaka University Hospital) during June 1994 through April 2000 and came to the follow-up clinics during July 2000 through December 2000 were studied. For LDR group, 26 oropharynx cancer patients who were treated by LDR Ir-192 interstitial brachytherapy with/without external beam radiotherapy in Nancy (Centre Alexis Vautrin) during February 1989 through July 1998 and came to the follow-up clinics during April 1999 through July 1999 were studied. The standard HDR schedules were 54 Gy/9 fr/5-6 days for monotherapy and 18-24 Gy/3-4 fr/2-3 days following 45 Gy external beam radiotherapy. The standard LDR schedules were 65 Gy/5-6 days for monotherapy and 15-25 Gy/2-3 days following 50 Gy external beam radiotherapy. For evaluation of the late changes, we scored the mucosal and muscular changes inside the treated volume using the modified Dische score system and the RTOG/EORTC late radiation morbidity scoring scheme. For 6 items of the modified Dische score system, no significant difference was found between HDR and LDR groups. For the remaining 2 items (pallor, mobility impairment of faucial pillars), LDR group showed higher scores (p=0.010, 0.002). LDR group showed a trend toward higher scores for the RTOG/EORTC scheme (p=0.059). Some predict late effects by HDR interstitial brachytherapy to be severer than by LDR because no dose-rate effects can be expected. Our study, however, showed at least equivalent or even milder late changes by HDR. Appropriate fractionation schedule and extra geometrical sparing effects by optimized dose distribution of HDR group might result in milder late changes. With our

  2. Phase III trial of high and low dose rate interstitial radiotherapy for early oral tongue cancer

    International Nuclear Information System (INIS)

    Inoue, Takehiro; Inoue, Toshihiko; Teshima, Teruki; Murayama, Shigeyuki; Shimizutani, Kimishige; Fuchihata, Hajime; Furukawa, Souhei

    1996-01-01

    Purpose: Oral tongue carcinomas are highly curable with radiotherapy. In the past, patients with tongue carcinoma have usually been treated with low dose rate (LDR) interstitial radiation. This Phase III study was designed to compare the treatment results obtained with LDR with those obtained with high dose rate (HDR) interstitial radiotherapy for tongue carcinoma. Methods and Materials: The criteria for patient selection for the Phase III study were: (a) presence of a T1T2N0 tumor that could be treated with single-plane implantation, (b) localization of tumor at the lateral tongue border, (c) tumor thickness of 10 mm or less, (d) performance status between O and 3, and (e) absence of any severe concurrent disease. From April 1992 through December 1993, 15 patients in the LDR group (70 Gy/4 to 9 days) and 14 patients in the HDR group (60 Gy/10 fractions/6 days) were accrued. The time interval between two fractions of the HDR brachytherapy was more than 6 h. Results: Local recurrence occurred in two patients treated with LDR brachytherapy but in none of the patients treated with HDR. One- and 2-year local control rates for patients in the LDR group were both 86%, compared with 100% in the HDR group (p = 0.157). There were four patients with nodal metastasis in the LDR group and three in the HDR group. Local recurrence occurred in two of the four patients with nodal metastases in the LDR group. One- and 2-year nodal control rates for patients in the LDR group are were 85%, compared with 79% in the HDR group. Conclusion: HDR fractionated interstitial brachytherapy can be an alternative to traditional LDR brachytherapy for early tongue cancer and eliminate the radiation exposure for medical staffs

  3. Interstitial and external radiotherapy in carcinoma of the soft palate and uvula

    International Nuclear Information System (INIS)

    Esche, B.A.; Haie, C.M.; Gerbaulet, A.P.; Eschwege, F.; Richard, J.M.; Chassagne, D.

    1988-01-01

    Forty-three patients, all male, with limited epidermoid carcinoma of the soft palate and uvula were treated by interstitial implant usually associated with external radiotherapy. Most patients received 50 Gy external irradiation to the oropharynx and neck followed by 20-35 Gy by interstitial iridium-192 wires using either guide gutters or a plastic tube technique. Twelve primary tumors and two recurrences after external irradiation alone had implant only for 65-75 Gy. Total actuarial local control is 92% with no local failures in 34 T1 primary tumors. Only one serious complication was seen. Overall actuarial survival was 60% at 3 years and 37% at 5 years but cause-specific survivals were 81% and 64%. The leading cause of death was other aerodigestive cancer, with an actuarial rate of occurrence of 10% per year after treatment of a soft palate cancer. Interstitital brachytherapy alone or combined with external irradiation is safe, effective management for early carcinoma of the soft palate and uvula but second malignancy is a serious problem

  4. [Brachytherapy of brainstem tumors].

    Science.gov (United States)

    Julow, Jenö; Viola, Arpád; Major, Tibor; Valálik, István; Sági, Sarolta; Mangel, László; Kovács, Rita Beáta; Repa, Imre; Bajzik, Gábor; Németh, György

    2004-01-20

    The optimal therapy of brain stem tumours of different histopathology determines the expected length of survival. Authors report 125Iodine interstitial irradiation of brain stem tumours with stereotactic brachytherapy. Two patients having brain stem tumours were suffering from glioma or from metastases of a carcinoma. In Case 1 the tumour volume was 1.98 cm3 at the time of planning interstitial irradiation. The control MRI examination performed at 42 months post-op showed a postirradiation cyst size of 5.73 cm3 indicating 65.5% shrinkage. In Case 2 the shrinkage was more apparent as the tumour volume measured on the control MRI at 8 months post-op was only 0.16 cm3 indicating 97.4% shrinkage of the 6.05 cm3 target volume at the time of brachytherapy with the metastasis practically disappearing. Quick access to histopathological results of the stereotactic intraoperative biopsy made it possible to carry out the 125Iodine stereotactic brachytherapy immediately after the biopsy, resulting in less inconvenience for patients of a second possible intervention. The control MRI scans show significant shrinkage of tumours in both patients. The procedure can be performed as a biopsy. The CT and image fusion guided 125Iodine stereotactic brachytherapy can be well planned dosimetrically and is surgically precise.

  5. WE-G-17A-05: Real-Time Catheter Localization Using An Active MR Tracker for Interstitial Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Wang, W; Damato, A; Viswanathan, A; Cormack, R [Dana Farber Cancer Institute / Brigham and Women' s Hospital, Boston, MA (United States); Harvard Medical School, Boston, MA (United States); Penzkofer, T; Schmidt, E [Brigham and Women' s Hospital, Boston, MA (United States); Pan, L; Gilson, W [Siemens Corporation, Corporate Technology, Baltimore, MD (United States); Seethamraju, R [Siemens Healthcare, Boston, MA (United States)

    2014-06-15

    Purpose: To develop a novel active MR-tracking system which can provide accurate and rapid localization of brachytherapy catheters, and assess its reliability and spatial accuracy in comparison to standard catheter digitization using MR images. Methods: An active MR tracker for brachytherapy was constructed by adding three printed-circuit micro-coils to the shaft of a commercial metallic stylet. A gel phantom with an embedded framework was built, into which fifteen 14-Gauge catheters were placed, following either with parallel or crossed paths. The tracker was inserted sequentially into each catheter, with MR-tracking running continuously. Tracking was also performed during the tracker's removal from each catheter. Catheter trajectories measured from the insertion and the removal procedures using the same micro-coil were compared, as well as trajectories obtained using different micro-coils. A 3D high-resolution MR image dataset of the phantom was acquired and imported into a treatment planning system (TPS) for catheter digitization. A comparison between MR-tracked positions and positions digitized from MR images by TPS was performed. Results: The MR tracking shows good consistency for varying catheter paths and for all micro-coils (mean difference ∼1.1 mm). The average distance between the MR-tracking trajectory and catheter digitization from the MR images was 1.1 mm. Ambiguity in catheter assignment from images due to crossed paths was resolved by active tracking. When tracking was interleaved with imaging, real-time images were continuously acquired at the instantaneous tip positions and displayed on an external workstation. Conclusion: The active MR tracker may be used to provide an independent measurement of catheter location in the MR environment, potentially eliminating the need for subsequent CT. It may also be used to control realtime imaging of catheter placement. This will enable MR-based brachytherapy planning of interstitial implants without

  6. WE-G-17A-05: Real-Time Catheter Localization Using An Active MR Tracker for Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Wang, W; Damato, A; Viswanathan, A; Cormack, R; Penzkofer, T; Schmidt, E; Pan, L; Gilson, W; Seethamraju, R

    2014-01-01

    Purpose: To develop a novel active MR-tracking system which can provide accurate and rapid localization of brachytherapy catheters, and assess its reliability and spatial accuracy in comparison to standard catheter digitization using MR images. Methods: An active MR tracker for brachytherapy was constructed by adding three printed-circuit micro-coils to the shaft of a commercial metallic stylet. A gel phantom with an embedded framework was built, into which fifteen 14-Gauge catheters were placed, following either with parallel or crossed paths. The tracker was inserted sequentially into each catheter, with MR-tracking running continuously. Tracking was also performed during the tracker's removal from each catheter. Catheter trajectories measured from the insertion and the removal procedures using the same micro-coil were compared, as well as trajectories obtained using different micro-coils. A 3D high-resolution MR image dataset of the phantom was acquired and imported into a treatment planning system (TPS) for catheter digitization. A comparison between MR-tracked positions and positions digitized from MR images by TPS was performed. Results: The MR tracking shows good consistency for varying catheter paths and for all micro-coils (mean difference ∼1.1 mm). The average distance between the MR-tracking trajectory and catheter digitization from the MR images was 1.1 mm. Ambiguity in catheter assignment from images due to crossed paths was resolved by active tracking. When tracking was interleaved with imaging, real-time images were continuously acquired at the instantaneous tip positions and displayed on an external workstation. Conclusion: The active MR tracker may be used to provide an independent measurement of catheter location in the MR environment, potentially eliminating the need for subsequent CT. It may also be used to control realtime imaging of catheter placement. This will enable MR-based brachytherapy planning of interstitial implants without

  7. INTERSTITIAL BRACHYTHERAPY USING TEMPLATE FOR LOCALLY ADVANCED GYNAECOLOGICAL MALIGNANCIES- REVISITING THE FORGOTTEN CLASSICAL ART- A SINGLE INSTITUTE EXPERIENCE

    Directory of Open Access Journals (Sweden)

    Chatharaju Swarna Kumari

    2017-10-01

    Full Text Available BACKGROUND Brachytherapy is an important therapeutic strategy for the treatment of locally advanced gynaecologic (GYN cancers despite evolution of different newer radiotherapy techniques like high-dose-rate and image-guided BT. Despite being used in the management of advanced gynaecological cancer, currently there is a scarcity of studies and data on interstitial BT in Indian context. This is a retrospective analysis on 71 patients with locally advanced gynaecological malignancies treated in the period of 2010 to 2016 to assess the local tumour control, survival, and complications with the template (Syed-Neblett guided interstitial technique. MATERIALS AND METHODS The patients with a median age of 51 years treated from July 2010 to May 2016 were retrospectively reviewed. This study included previously unirradiated 71 patients with advance stage of gynaecological malignancies, not suitable for intracavitary brachytherapy due to distorted anatomy or extensive disease stage. Histologically all patients had squamous cell carcinoma (cervix= 56, vault= 9, vagina= 6 and treated by whole pelvis external beam radiation therapy (EBRT up to a total dose of 50 Gy in 25 fractions. These patients were further treated by high-dose-rate interstitial brachytherapy using SyedNeblett dedicated vaginal plastic template. During treatment all these patients were re-optimised and a dose of 15-21 Gy was delivered in 3 fractions with a minimum gap of 6 hours between fractions using Varisource iX HDR unit. RESULTS Out of 71 patients 5 were lost to followup during study period and they were excluded from the final analysis. The average followup duration ranged between 6-71 months and median followup was 20 months. This study included parameters like local disease control, acute/late complications and distant metastasis. Out of 66 patients, local disease control was seen in 54 (81.81% patients, whereas local recurrence was observed in 12 patients (18.18%. Distant

  8. Limb sparing surgery and boost with high dose rate interstitial brachytherapy in treatment of soft tissue sarcoma

    International Nuclear Information System (INIS)

    Koike, P.; Miziara, M.; Soares, C.; Fogaroli, R.; Baraldi, H.; Pellizoni, A.; Borba, G.

    2003-01-01

    Soft tissue sarcoma, a rare neoplasia with high aggressively, accounts for approximately 0,7% of the malignant tumors and occurs more often with youngest. Because of the potential risk of local recurrence, theoretically surgical resection, encompassing macroscopic tumor with a margin of macroscopically noninvolved tissue is the right way to perform, with a wide 'en bloc' resection, amputation, with bad functional results. The aim of conservative treatment is combined modality therapy as surgical resection and irradiation to obtain a local control rate as high as possible while preserving functional results. A retrospective review of 31 patients treated with high dose rate (HDR) Brachytherapy in the Radiotherapy Department Arnaldo Vieira de Carvalho Cancer Institute. Methods: Between April 1995 and August 1999, 31 patients who underwent a combine therapy; examined the results on multivariate analysis of conservative surgery and brachytherapy follow/or not by external beam radiation (EBRT). The 31 patients treated, 17 ( 54,8%) females and 14(45,2%) males have a median age of 48 years ( range,19 to 77 years). Most of the tumors was located in the lower limb (17/31 - 54,8%) . The other sites were the upper limb (10/31-32,3%), thoracic wall and abdomen (3/31-9,7%).Classification of the tumors, according to the International Union Against Cancer (UICC) staging was T1 5 patients (16%), T2 (24/31-77%). Median size of the tumors was 9,2cm ( ranged, 2,5 to 24cm). Most of the tumors being malignant fibrous histiocytomas (9/31-29%) and the histological grade II (14/31-45%). Twenty-two (71%) patients had intraoperative implants and the insertion of the radioactive source was delayed 24 to 120 hours. Eight patients (25,8%) had postoperative and received HDRB 45 to 60 days after the surgery . Guide needles were placed in the tumor bed, perpendicular to the scar, systematically in a single plane, the implant volume being defined by radiotherapist . A minimum safety margin of 2 cm

  9. Interim report of image-guided conformal high-dose-rate brachytherapy for patients with unfavorable prostate cancer: the William Beaumont Phase II dose-escalating trial

    International Nuclear Information System (INIS)

    Martinez, Alvaro A.; Kestin, Larry L.; Stromberg, Jannifer S.; Gonzalez, Jose A.; Wallace, Michelle; Gustafson, Gary S.; Edmundson, Gregory K.; Spencer, William; Vicini, Frank A.

    2000-01-01

    Purpose: We analyzed our institution's experience treating patients with unfavorable prostate cancer in a prospective Phase II dose-escalating trial of external beam radiation therapy (EBRT) integrated with conformal high-dose-rate (HDR) brachytherapy boosts. This interim report discusses treatment outcome and prognostic factors using this treatment approach. Methods and Materials: From November 1991 through February 1998, 142 patients with unfavorable prostate cancer were prospectively treated in a dose-escalating trial with pelvic EBRT in combination with outpatient HDR brachytherapy at William Beaumont Hospital. Patients with any of the following characteristics were eligible: pretreatment prostate-specific antigen (PSA) ≥ 10.0 ng/ml, Gleason score ≥ 7, or clinical stage T2b or higher. All patients received pelvic EBRT to a median total dose of 46.0 Gy. Pelvic EBRT was integrated with ultrasound-guided transperineal conformal interstitial iridium-192 HDR implants. From 1991 to 1995, 58 patients underwent three conformal interstitial HDR implants during the first, second, and third weeks of pelvic EBRT. After October 1995, 84 patients received two interstitial implants during the first and third weeks of pelvic EBRT. The dose delivered via interstitial brachytherapy was escalated from 5.50 Gy to 6.50 Gy for each implant in those patients receiving three implants, and subsequently, from 8.25 Gy to 9.50 Gy per fraction in those patients receiving two implants. To improve implant quality and reduce operator dependency, an on-line, image-guided interactive dose optimization program was utilized during each HDR implant. No patient received hormonal therapy unless treatment failure was documented. The median follow-up was 2.1 years (range: 0.2-7.2 years). Biochemical failure was defined according to the American Society for Therapeutic Radiology and Oncology Consensus Panel definition. Results: The pretreatment PSA level was ≥ 10.0 ng/ml in 51% of patients. The

  10. Comparison of prostate contours between conventional stepping transverse imaging and Twister-based sagittal imaging in permanent interstitial prostate brachytherapy.

    Science.gov (United States)

    Kawakami, Shogo; Ishiyama, Hiromichi; Satoh, Takefumi; Tsumura, Hideyasu; Sekiguchi, Akane; Takenaka, Kouji; Tabata, Ken-Ichi; Iwamura, Masatsugu; Hayakawa, Kazushige

    2017-08-01

    To compare prostate contours on conventional stepping transverse image acquisitions with those on twister-based sagittal image acquisitions. Twenty prostate cancer patients who were planned to have permanent interstitial prostate brachytherapy were prospectively accrued. A transrectal ultrasonography probe was inserted, with the patient in lithotomy position. Transverse images were obtained with stepping movement of the transverse transducer. In the same patient, sagittal images were also obtained through rotation of the sagittal transducer using the "Twister" mode. The differences of prostate size among the two types of image acquisitions were compared. The relationships among the difference of the two types of image acquisitions, dose-volume histogram (DVH) parameters on the post-implant computed tomography (CT) analysis, as well as other factors were analyzed. The sagittal image acquisitions showed a larger prostate size compared to the transverse image acquisitions especially in the anterior-posterior (AP) direction ( p transverse image acquisitions was correlated to DVH parameters such as D 90 ( R = 0.518, p = 0.019), and V 100 ( R = 0.598, p = 0.005). There were small but significant differences in the prostate contours between the transverse and the sagittal planning image acquisitions. Furthermore, our study suggested that the differences between the two types of image acquisitions might correlated to dosimetric results on CT analysis.

  11. Matched-pair analysis and dosimetric variations of two types of software for interstitial permanent brachytherapy for prostate cancer

    Energy Technology Data Exchange (ETDEWEB)

    Ishiyama, Hiromichi, E-mail: hishiyam@kitasato-u.ac.jp [Department of Radiology, Kitasato University School of Medicine, Sagamihara, Kanagawa (Japan); Nakamura, Ryuji [Department of Radiology, Iwate Medical University, Morioka, Iwate (Japan); Satoh, Takefumi [Department of Urology, Kitasato University School of Medicine, Sagamihara, Kanagawa (Japan); Tanji, Susumu [Department of Urology, Iwate Medical University, Morioka, Iwate (Japan); Teh, Bin S. [Department of Radiation Oncology, The Methodist Hospital, Houston, TX (United States); Uemae, Mineko [Division of Radiation Oncology, Kitasato University Hospital, Sagamihara, Kanagawa (Japan); Baba, Shiro [Department of Urology, Kitasato University School of Medicine, Sagamihara, Kanagawa (Japan); Hayakawa, Kazushige [Department of Radiology, Kitasato University School of Medicine, Sagamihara, Kanagawa (Japan)

    2012-04-01

    The purpose of this study was to determine whether identical dosimetric results could be achieved using different planning software for permanent interstitial brachytherapy for prostate cancer. Data from 492 patients treated with brachytherapy were used for matched-pair analysis. Interplant and Variseed were used as software for ultrasound-based treatment planning. Institution, neoadjuvant hormonal therapy, prostate volume, and source strength were used for factors to match the 2 groups. The study population comprised of 126 patients with treatment planning using Interplant software and 127 matched patients using Variseed software. Dosimetric results were compared between the 2 groups. The Variseed group showed significantly higher values for dose covering 90% of prostate volume (pD90), prostate volume covered by 150% of prescription dose (pV150), and dose covering 30% of the urethra (uD30) compared with the Interplant group. Our results showed that use of different software could lead to different dosimetric results, which might affect the clinical outcomes.

  12. Interstitial vaginal needle implantation in gynecological tumors : design and construction of applicator

    International Nuclear Information System (INIS)

    Kang, Seung Hee; Chun, Mi Son; Kang, Hae Jin; Jung, Chil; Son, Jeong Hyae

    1998-01-01

    It is not a simple task to achieve the ideal isodose curve with a standard vaginal applicator or single plane needle impant in the paravaginal tissue when primary or recurrent gynecological neoplasms(cervical cancers, vaginal cancers and vulvar cancers) are treated as a boost following external beam radiotherapy. The authors introduce the development and construction of a simple, inexpensive, customized applicator for volume implant to maximize the radiation dose to the tumor while minimizing the dose to the rectum and the bladder. Nine patients underwent Ir-192 transperineal interstitial implantation for either recurrent(5 cases)or primary(3 cases) cervical cancers or primary vaginal cancer(1 case) between August 1994 and February 1998 at Ajou university hospital. First 3 cases were performed with a single plane implant guided by digital palpation. Because of inadequate isodose coverage in the tumor volume in first 3 cases, we designed and constructed interstitial vaginal applicator for volume implant to mprove tumor dose distribution and homogeneity while sparing the surrounding normal tissue. Our applicators consist of vaginal obturator and perineal template that made of the clear acrylamide and dental mold material(Provil). The applicators were customized individually according to the tumor size and its location. Both HDR and LDR irradiation were given with these applicators accomodating 6 Fr needles(Microselectron, Nucletron). The pretreatment planning prior to actual implant was performed whenever possible. Needles can be inserted easily and evenly into the tumor volume through the holes of templates, requiring less efforts and time for the implant prodecure. Our applicators made of materials available from commercial vendors. These have an advantage that require easy procedure, and spend relatively short time to construct. Also it was possible to fabricate applicators to individualize according to the tumor size and its location and to achieve the ideal

  13. Permanent Seed Implant Dosimetry (PSID)TM 4.5 version as isodose and Treatment Planning System (TPS) programme for brachytherapy

    International Nuclear Information System (INIS)

    Indra Saptiama; Moch Subechi; Anung Pujiyanto; Hotman Lubis; Herlan Setiawan

    2014-01-01

    The medical treatment using radiation therapy for cancer diseases is increasingly developed. One of the method used in radiotherapy is brachytherapy. Brachytherapy is radiation therapy method in which a radiation source is implanted in cancer cell directly so the dose accepted by cancer cell is the highest dose and the dose accepted by normal cell is the lowest dose. I-125 Seed have been made successfully in domestic. To support the implant of I-125 seed for brachytherapy needs computer programme for the isodose calculation and Treatment Planning System (TPS). Permanent Seed Implant Dosimetry (PSID) 4.5 is one of the isodose calculation and Treatment Planning System (TPS) programme that is owned by Center for Radioisotope and Radiopharmaceutical-BATAN. In isodose calculation, PSID 4.5 uses 1D formalism and 2D formalism based on AAPM-TG43 (Association of American Physicist in Medicine- Task Group No.43). Anisotropic function on 1D formalism depend on distance function while on 2D formalism count on distance and angle function therefore 2D formalism has isodose calculation better than 1D formalism usage. PSID 4.5 can display the isodose contour of the seed I-125 radiation source in 2 dimension (2D) and 3 dimension (3D). The computer programme of isodose calculation and TPS uses PSID 4.5 is expected able to help planning for seed I-125 implantation process for brachytherapy that used by paramedics and to support the usage of seed I-125 as domestic product. (author)

  14. Radioactive seed immobilization techniques for interstitial brachytherapy

    International Nuclear Information System (INIS)

    Yan, K.; Podder, T.; Buzurovic, I.; Hu, Y.; Dicker, A.; Valicenti, R.; Yu, Y.; Messing, E.; Rubens, D.; Sarkar, N.; Ng, W.

    2008-01-01

    In prostate brachytherapy, seeds can detach from their deposited sites and move locally in the pelvis or migrate to distant sites including the pulmonary and cardiac regions. Undesirable consequences of seed migration include inadequate dose coverage of the prostate and tissue irradiation effects at the site of migration. Thus, it is clinically important to develop seed immobilization techniques. We first analyze the possible causes for seed movement, and propose three potential techniques for seed immobilization: (1) surgical glue, (2) laser coagulation and (3) diathermy coagulation. The feasibility of each method is explored. Experiments were carried out using fresh bovine livers to investigate the efficacy of seed immobilization using surgical glue. Results have shown that the surgical glue can effectively immobilize the seeds. Evaluation of the radiation dose distribution revealed that the non-immobilized seed movement would change the planned isodose distribution considerably; while by using surgical glue method to immobilize the seeds, the changes were negligible. Prostate brachytherapy seed immobilization is necessary and three alternative mechanisms are promising for addressing this issue. Experiments for exploring the efficacy of the other two proposed methods are ongoing. Devices compatible with the brachytherapy procedure will be designed in future. (orig.)

  15. Distant Metastases Following Permanent Interstitial Brachytherapy for Patients With Clinically Localized Prostate Cancer

    International Nuclear Information System (INIS)

    Taira, Al V.; Merrick, Gregory S.; Galbreath, Robert W.; Butler, Wayne M.; Lief, Jonathan; Adamovich, Edward; Wallner, Kent E.

    2012-01-01

    Purpose: Recent publications have suggested high-risk patients undergoing radical prostatectomy have a lower risk of distant metastases and improved cause-specific survival (CSS) than patients receiving definitive external beam radiation therapy (XRT). To date, none of these studies has compared distant metastases and CSS in brachytherapy patients. In this study, we evaluate such parameters in a consecutive cohort of brachytherapy patients. Methods and Materials: From April 1995 to June 2007, 1,840 consecutive patients with clinically localized prostate cancer were treated with brachytherapy. Risk groups were stratified according to National Comprehensive Cancer Network ( (www.nccn.org)) guidelines. Subgroups of 658, 893, and 289 patients were assigned to low, intermediate, and high-risk categories. Median follow-up was 7.2 years. Along with brachytherapy implantation, 901 (49.0%) patients received supplemental XRT, and 670 (36.4%) patients received androgen deprivation therapy (median duration, 4 months). The mode of failure (biochemical, local, or distant) was determined for each patient for whom therapy failed. Cause of death was determined for each deceased patient. Multiple parameters were evaluated for impact on outcome. Results: For the entire cohort, metastases-free survival (MFS) and CSS at 12 years were 98.1% and 98.2%, respectively. When rates were stratified by low, intermediate, and high-risk groups, the 12-year MFS was 99.8%, 98.1%, and 93.8% (p < 0.001), respectively. CSS rates were 99.8%, 98.0%, and 95.3% (p < 0.001) for low, intermediate, and high-risk groups, respectively. Biochemical progression-free survival was 98.7%, 95.9% and 90.4% for low, intermediate, and high-risk patients, respectively (p < 0.001). In multivariate Cox-regression analysis, MFS was mostly closely related to Gleason score and year of treatment, whereas CSS was most closely associated with Gleason score. Conclusions: Excellent CSS and MFS rates are achievable with high

  16. Dose optimization in simulated permanent interstitial implant of prostate brachytherapy

    International Nuclear Information System (INIS)

    Faria, Fernando Pereira de

    2006-01-01

    Any treatment of cancer that uses some modality of radiotherapy is planned before being executed. In general the goal in radiotherapy is to irradiate the target to be treated minimizing the incidence of radiation in healthy surrounding tissues. The planning differ among themselves according to the modality of radiotherapy, the type of cancer and where it is located. This work approaches the problem of dose optimization for the planning of prostate cancer treatment through the modality of low dose-rate brachytherapy with Iodine 125 or Palladium 103 seeds. An algorithm for dose calculation and optimization was constructed to find the seeds configuration that better fits the relevant clinical criteria such as as the tolerated dose by the urethra and rectum and the desired dose for prostate. The algorithm automatically finds this configuration from the prostate geometry established in two or three dimensions by using images of ultrasound, magnetic resonance or tomography and from the establishment of minimum restrictions to the positions of the seeds in the prostate and needles in a template. Six patterns of seeds distribution based on clinical criteria were suggested and tested in this work. Each one of these patterns generated a space of possible seeds configurations for the prostate tested by the dose calculation and optimization algorithm. The configurations that satisfied the clinical criteria were submitted to a test according to an optimization function suggested in this work. The configuration that produced maximum value for this function was considered the optimized one. (author)

  17. A Dose-Volume Analysis of Magnetic Resonance Imaging-Aided High-Dose-Rate Image-Based Interstitial Brachytherapy for Uterine Cervical Cancer

    International Nuclear Information System (INIS)

    Yoshida, Ken; Yamazaki, Hideya; Takenaka, Tadashi; Kotsuma, Tadayuki; Yoshida, Mineo; Furuya, Seiichi; Tanaka, Eiichi; Uegaki, Tadaaki; Kuriyama, Keiko; Matsumoto, Hisanobu; Yamada, Shigetoshi; Ban, Chiaki

    2010-01-01

    Purpose: To investigate the feasibility of our novel image-based high-dose-rate interstitial brachytherapy (HDR-ISBT) for uterine cervical cancer, we evaluated the dose-volume histogram (DVH) according to the recommendations of the Gynecological GEC-ESTRO Working Group for image-based intracavitary brachytherapy (ICBT). Methods and Materials: Between June 2005 and June 2007, 18 previously untreated cervical cancer patients were enrolled. We implanted magnetic resonance imaging (MRI)-available plastic applicators by our unique ambulatory technique. Total treatment doses were 30-36 Gy (6 Gy per fraction) combined with external beam radiotherapy (EBRT). Treatment plans were created based on planning computed tomography with MRI as a reference. DVHs of the high-risk clinical target volume (HR CTV), intermediate-risk CTV (IR CTV), and the bladder and rectum were calculated. Dose values were biologically normalized to equivalent doses in 2-Gy fractions (EQD 2 ). Results: The median D90 (HR CTV) and D90 (IR CTV) per fraction were 6.8 Gy (range, 5.5-7.5) and 5.4 Gy (range, 4.2-6.3), respectively. The median V100 (HR CTV) and V100 (IR CTV) were 98.4% (range, 83-100) and 81.8% (range, 64-93.8), respectively. When the dose of EBRT was added, the median D90 and D100 of HR CTV were 80.6 Gy (range, 65.5-96.6) and 62.4 Gy (range, 49-83.2). The D 2cc of the bladder was 62 Gy (range, 51.4-89) and of the rectum was 65.9 Gy (range, 48.9-76). Conclusions: Although the targets were advanced and difficult to treat effectively by ICBT, MRI-aided image-based ISBT showed favorable results for CTV and organs at risk compared with previously reported image-based ICBT results.

  18. A dose-volume analysis of magnetic resonance imaging-aided high-dose-rate image-based interstitial brachytherapy for uterine cervical cancer.

    Science.gov (United States)

    Yoshida, Ken; Yamazaki, Hideya; Takenaka, Tadashi; Kotsuma, Tadayuki; Yoshida, Mineo; Furuya, Seiichi; Tanaka, Eiichi; Uegaki, Tadaaki; Kuriyama, Keiko; Matsumoto, Hisanobu; Yamada, Shigetoshi; Ban, Chiaki

    2010-07-01

    To investigate the feasibility of our novel image-based high-dose-rate interstitial brachytherapy (HDR-ISBT) for uterine cervical cancer, we evaluated the dose-volume histogram (DVH) according to the recommendations of the Gynecological GEC-ESTRO Working Group for image-based intracavitary brachytherapy (ICBT). Between June 2005 and June 2007, 18 previously untreated cervical cancer patients were enrolled. We implanted magnetic resonance imaging (MRI)-available plastic applicators by our unique ambulatory technique. Total treatment doses were 30-36 Gy (6 Gy per fraction) combined with external beam radiotherapy (EBRT). Treatment plans were created based on planning computed tomography with MRI as a reference. DVHs of the high-risk clinical target volume (HR CTV), intermediate-risk CTV (IR CTV), and the bladder and rectum were calculated. Dose values were biologically normalized to equivalent doses in 2-Gy fractions (EQD(2)). The median D90 (HR CTV) and D90 (IR CTV) per fraction were 6.8 Gy (range, 5.5-7.5) and 5.4 Gy (range, 4.2-6.3), respectively. The median V100 (HR CTV) and V100 (IR CTV) were 98.4% (range, 83-100) and 81.8% (range, 64-93.8), respectively. When the dose of EBRT was added, the median D90 and D100 of HR CTV were 80.6 Gy (range, 65.5-96.6) and 62.4 Gy (range, 49-83.2). The D(2cc) of the bladder was 62 Gy (range, 51.4-89) and of the rectum was 65.9 Gy (range, 48.9-76). Although the targets were advanced and difficult to treat effectively by ICBT, MRI-aided image-based ISBT showed favorable results for CTV and organs at risk compared with previously reported image-based ICBT results. (c) 2010 Elsevier Inc. All rights reserved.

  19. Needle displacement during HDR brachytherapy in the treatment of prostate cancer

    International Nuclear Information System (INIS)

    Damore, Steven J.; Syed, A.M. Nisar; Puthawala, Ajmel A.; Sharma, Anil

    2000-01-01

    Purpose: We used clinical patient data to examine implant displacement between high dose rate (HDR) brachytherapy fractions for prostate cancer to determine its impact on treatment delivery. Materials and Methods: We analyzed the verification films taken prior to each fraction for 96 consecutive patients treated with HDR brachytherapy boosts as part of their radiation therapy for definitive treatment of organ-confined prostate cancer at our institution. Patients were treated with 18-24 Gy in 4 fractions of HDR delivered in 40 hours followed by 36-39.6 Gy external beam radiation to the prostate. We determined the mean and maximum displacement distances of marker seeds placed in the prostate and of the implanted needles between HDR fractions. Results: Mean and maximum displacement distances between fractions were documented up to 7.6 mm and 28.5 mm, respectively, for the implant needles and 3.6 mm and 11.4 mm, respectively, for the gold marker seeds. All displacement of implant needles occurred in the caudal direction. At least 1 cm caudal displacement of needles occurred prior to 15.5% all fractions. Manual adjustment of needles was required prior to 15% of fractions, and adjustment of the CLP only was required in 24%. Most of the displacement for both the marker seeds and needles occurred between the first and second fractions. Conclusions: There is significant caudal displacement of interstitial implant needles between HDR fractions in our prostate cancer patients. Obtaining verification films and making adjustments in the treatment volume prior to each fraction is necessary to avoid significant inaccuracies in treatment delivery

  20. Comparison between external beam radiotherapy (70 Gy/74 Gy) and permanent interstitial brachytherapy in 890 intermediate risk prostate cancer patients

    International Nuclear Information System (INIS)

    Goldner, Gregor; Pötter, Richard; Battermann, Jan J.; Kirisits, Christian; Schmid, Maximilian P.; Sljivic, Samir; Vulpen, Marco van

    2012-01-01

    Purpose: Aim of this analysis was to compare biochemical no evidence of disease (bNED) rates in intermediate-risk prostate-cancer patients treated at two centres of excellence using different approaches: permanent interstitial brachytherapy (BT) and external beam radiotherapy (EBRT). Materials and methods: A total of 890 intermediate-risk prostate-cancer patients, who were treated from 1998 to 2008, were identified in the two local databases. In Utrecht 601 patients received I-125 BT applying a dose of 144 Gy. In Vienna 289 patients were treated by EBRT, applying a local dose of 70 Gy in 105 patients and 74 Gy in 184 patients. bNED-rates (Phoenix-definition) were assessed. Results: Median follow-up was 48 months (1–150). 5-Year actuarial bNED-rates were 81% for BT-patients and 75% for EBRT-patients (67% for 70 Gy and 82% for 74 Gy), respectively. In univariate analysis no difference between BT and EBRT could be detected. In multivariate analysis including tumour-stage, GleasonScore, initial PSA, hormonal therapy and treatment-centre (BT vs. EBRT) only T-stage, GleasonScore and PSA were found to be significant. Additional analysis including radiation dose showed the same outcome. Conclusions: Intermediate-risk prostate cancer patients treated by permanent interstitial brachytherapy show biochemical tumour-control-rates which are comparable to EBRT of 74 Gy.

  1. A simple device to protect against osteoradionecrosis induced by interstitial irradiation

    International Nuclear Information System (INIS)

    Levendag, P.C.; Visch, L.L.; Driver, N.

    1990-01-01

    The incidence of osteoradionecrosis has declined since the introduction of preventive oral hygiene programs and meticulous dental evaluations before and after irradiation. Nevertheless, radiation dose per se still remains an important factor in osteoradionecrosis. Interstitial radiation has received much attention in the past decade since the use of flexible afterloading systems. It has become common practice in large oncology centers to implant radiation carriers in bulky tumor in the oral cavity and/or oropharynx. For interstitial radiation, with or without external radiation, minimal tumor doses are often cited to be more than 70 Gy. Unfortunately, if the mandible receives more than 70 Gy, it is at risk for the development of osteoradionecrosis. Therefore a simple protective lead device has been designed for routine use in brachytherapy in oral cavity tumors to reduce the dose to the mandible. This device will diminish the potential risk of osteoradionecrosis development

  2. CT stereotactic reconstruction of oral cavity interstitial plastic tube implants

    International Nuclear Information System (INIS)

    Crispin, V.; Carrasco, P.; Guardino, C.; Lopez, J.; Chust, M.; Arribas, L.; Mengual, J.; Miragall, E.G.; Hernandez, A.; Carrascosa, M.; Cardenal, R.; Guinot, J.; Casana, M.; Prats, C.

    1996-01-01

    The continuous using of CT images in external RT have made us think of its applications for plastic tube interstitial implants in the oral cavity in order to calculate the dose delivered by an interstitial implant at any point of the image and its relationship with local control and complications. Moreover, the outcoming result of the whole treatment depends on whether the irradiated volume up to a prescribed dose includes the CTV or not. None of these objectives may be achieved through the classical film reconstruction. Although film reconstruction appeared as the only accurate method for these purposes in the early eighties, it does not allow us to calculate doses at critical points or volumes. Therefore possible complications over critical tissues surrounding the radioactive implant cannot be taken into account in a precise way. The use of a stereotactic coordinate system could make CT reconstruction as precise as film reconstruction. As our stereotactic frame can be placed over the patient in 'direct' or 'inverse' positions it is really interesting in the applications we are talking about. We also have used a non invasive standard plexiglass helmet commonly used in stereotactic fractionated irradiations in teletherapy. It fits perfectly the patient's head and avoids any movement of the patient during the CT exam. We do parallel slices, approximately perpendicular to the iridium wires (following the Paris System), covering the whole implant helping ourselves with both bone and implant references. The dose-volume histograms and DNR (dose nonuniformity ratio) index defined by Saw et Al are used for intercomparison between the ortogonal and the stereotactic reconstructions. The existence of a minimum in the DNR curve indicates that there is a reference dose rate for this implant which provides an optimal dose distribution. If we calculate which is the minimum of each method, we find they are very close. So, as both methods give very similar results, we can conclude

  3. Edema worsens target coverage in high-dose-rate interstitial brachytherapy of mobile tongue cancer: a report of two cases.

    Science.gov (United States)

    Yoshida, Ken; Yamazaki, Hideya; Kotsuma, Tadayuki; Akiyama, Hironori; Takenaka, Tadashi; Masui, Koji; Yoshioka, Yasuo; Uesugi, Yasuo; Shimbo, Taiju; Yoshikawa, Nobuhiko; Yoshioka, Hiroto; Arika, Takumi; Tanaka, Eiichi; Narumi, Yoshifumi

    2017-02-01

    We report our study on two patients to highlight the risk of underdosage of the clinical target volume (CTV) due to edema during high-dose-rate interstitial brachytherapy (HDR-ISBT) of mobile tongue cancer. To treat the lateral side of the CTV, flexible applicator tubes were implanted on the mouth floor. Two-dimensional planning was performed using X-ray images for Case 1, and three-dimensional (3D) planning was performed using computed tomography (CT) for Case 2. Prescribed doses for both cases were 54 Gy in nine fractions. Case 1 was treated for cancer of the right lateral border of the tongue in 2005. Tongue edema occurred after implantation, and part of the lateral border of the tongue protruded between the applicator tubes. Acute mucosal reaction abated in the protruded area earlier than in the other parts of the CTV. In this case, the tumor recurred in this area 5 months after the treatment. Case 2 was treated for cancer of the left lateral border of the tongue. Because tongue edema occurred in this case also, plastic splints were inserted between the applicator tubes to push the edematous region into the irradiated area. The mucosal surface of the CTV was covered by the 70% isodose, and 100% isodose line for before and after splint insertion. Local control of the tumor was achieved 4 years after treatment. To ensure sufficient target coverage, 3D image-based planning using CT should be performed, followed by re-planning using repeated CT as needed. Also, the development of devices to prevent protrusion of the edematous tissue outside the target area will help to ensure the full dosing of CTV.

  4. CT-guided brachytherapy. A novel percutaneous technique for interstitial ablation of liver malignancies; CT-gesteuerte Brachytherapie. Eine neue perkutane Technik zur interstitiellen Ablation von Lebermetastasen

    Energy Technology Data Exchange (ETDEWEB)

    Ricke, J.; Wust, P.; Stohlmann, A.; Beck, A.; Cho, C.H.; Pech, M.; Wieners, G.; Spors, B.; Werk, M.; Rosner, C.; Haenninen, E.L.; Felix, R. [Klinik fuer Strahlenheilkunde, Charite Virchow-Klinikum, Humboldt-Univ. zu Berlin (Germany)

    2004-05-01

    Purpose: to assess safety and efficacy of CT-guided brachytherapy of liver malignancies. Patients and methods: 21 patients with 21 liver malignancies (19 metastases, two primary liver tumors) were treated with interstitial CT-guided brachytherapy applying a {sup 192}Ir source. In all patients, the use of image-guided thermal tumor ablation such as by radiofrequency or laser-induced thermotherapy (LITT) was impeded either by tumor size {>=} 5 cm in seven, adjacent portal or hepatic vein in ten, or adjacent bile duct bifurcation in four patients. Dosimetry was performed using three-dimensional CT data sets acquired after CT-guided positioning of the brachytherapy catheters. Results: the mean tumor diameter was 4.6 cm (2.5-11 cm). The mean minimal tumor dose inside the tumor margin amounted to 17 Gy (12-20 Gy). The proportion of the liver parenchyma exposed to > 5 gy was 18% (5-39%) of total liver parenchyma minus tumor volume. Nausea and vomiting were observed in six patients after brachytherapy (28%). One patient demonstrated obstructive jaundice due to tumor edema after irradiation of a metastasis adjacent to the bile duct bifurcation. We commonly encountered asymptomatic increases of liver enzymes. Local control rates after 6 and 12 months were 87% and 70%, respectively. Conclusion: CT-guided brachytherapy is safe and effective. This technique displays broader indications compared to image-guided thermal ablation by radiofrequency or LITT with respect to tumor size or localization. (orig.) [German] Ziel: Analyse der Sicherheit und Effektivitaet CT-gesteuerter Brachytherapie zur Ablation von Lebermalignomen. Patienten und Methodik: 21 Patienten mit 21 Lebermalignomen (19 Metastasen, zwei primaere Lebermalignome) wurden mit perkutaner, CT-gesteuerter interstitieller Brachytherapie mit {sup 192}Ir behandelt. Alle Patienten wiesen Umstaende auf, die eine bildgefuehrte thermische Ablation mit Radiofrequenz oder laserinduzierter Thermotherapie (LITT) einschraenkten

  5. Dosimetric evaluation of a novel high dose rate (HDR) intraluminal / interstitial brachytherapy applicator for gastrointestinal and bladder cancers

    Science.gov (United States)

    Aghamiri, Seyyed Mahmoud Reza; Najarian, Siamak; Jaberi, Ramin

    2010-01-01

    High dose rate (HDR) brachytherapy is one of the accepted treatment modalities in gastro‐intestinal tract and bladder carcinomas. Considering the shortcoming of contact brachytherapy routinely used in gastrointestinal tract in treatment of big tumors or invasive method of bladder treatment, an intraluminal applicator with the capability of insertion into the tumor depth seems to be useful. This study presents some dosimetric evaluations to introduce this applicator to the clinical use. The radiation attenuation characteristics of the applicator were evaluated by means of two dosimetric methods including well‐type chamber and radiochromic film. The proposed 110 cm long applicator has a flexible structure made of stainless steel for easy passage through lumens and a needle tip to drill into big tumors. The 2 mm diameter of the applicator is thick enough for source transition, while easy passage through any narrow lumen such as endoscope or cystoscope working channel is ensured. Well‐chamber results showed an acceptably low attenuation of this steel springy applicator. Performing absolute dosimetry resulted in a correlation coefficient of R=0.9916(p‐value≈10−7) between standard interstitial applicator and the one proposed in this article. This study not only introduces a novel applicator with acceptable attenuation but also proves the response independency of the GAFCHROMIC EBT films to energy. By applying the dose response of the applicator in the treatment planning software, it can be used as a new intraluminal / interstitial applicator. PACS number: 87.53.Bn, 87.53.Jw, 29.40.Cs

  6. Prostate cancer brachytherapy

    International Nuclear Information System (INIS)

    Abreu, Carlos Eduardo Vita; Silva, Joao L. F.; Srougi, Miguel; Nesrallah, Adriano

    1999-01-01

    The transperineal brachytherapy with 125 I/Pd 103 seed implantation guided by transurethral ultrasound must be presented as therapeutical option of low urinary morbidity in patients with localized prostate cancer. The combined clinical staging - including Gleason and initial PSA - must be encouraged, for definition of a group of low risk and indication of exclusive brachytherapy. Random prospective studies are necessary in order to define the best role of brachytherapy, surgery and external beam radiation therapy

  7. Objective and Longitudinal Assessment of Dermatitis After Postoperative Accelerated Partial Breast Irradiation Using High-Dose-Rate Interstitial Brachytherapy in Patients With Breast Cancer Treated With Breast Conserving Therapy: Reduction of Moisture Deterioration by APBI

    International Nuclear Information System (INIS)

    Tanaka, Eiichi; Yamazaki, Hideya; Yoshida, Ken; Takenaka, Tadashi; Masuda, Norikazu; Kotsuma, Tadayuki; Yoshioka, Yasuo; Inoue, Takehiro

    2011-01-01

    Purpose: To objectively evaluate the radiation dermatitis caused by accelerated partial breast irradiation (APBI) using high-dose-rate interstitial brachytherapy. Patients and Methods: The skin color and moisture changes were examined using a newly installed spectrophotometer and corneometer in 22 patients who had undergone APBI using open cavity implant high-dose-rate interstitial brachytherapy (36 Gy in six fractions) and compared with the corresponding values for 44 patients in an external beam radiotherapy (EBRT) control group (50–60 Gy in 25–30 fractions within 5–6 weeks) after breast conserving surgery. Results: All values changed significantly as a result of APBI. The extent of elevation in a∗ (reddish) and reduction in L∗ (black) values caused by APBI were similar to those for EBRT, with slightly delayed recovery for 6–12 months after treatment owing to the surgical procedure. In contrast, only APBI caused a change in the b∗ values, and EBRT did not, demonstrating that the reduction in b∗ values (yellowish) depends largely on the surgical procedure. The changes in moisture were less severe after APBI than after EBRT, and the recovery was more rapid. The toxicity assessment using the Common Toxicity Criteria, version 3, showed that all dermatitis caused by APBI was Grade 2 or less. Conclusion: An objective analysis can quantify the effects of APBI procedures on color and moisture cosmesis. The radiation dermatitis caused by APBI using the present schedule showed an equivalent effect on skin color and a less severe effect on moisture than the effects caused by standard EBRT.

  8. Ten cases of metastatic cervical cancer with the treatment of permanent 125I seeds interstitial implants

    International Nuclear Information System (INIS)

    Zhang Hongwei; Li Naibin; Li Qingxin; Liu Huiping; Meng Hui; Chao Dong

    2011-01-01

    Objective: To investigate the clinical effect of permanent 125 I seeds interstitial implants for metastatic cervical cancer. Methods: Under the guidance of the B-sonography, 125 I seeds were implanted into the eleven cervical lymph nodes of ten patients who had been given tumor resection. The pain relief and tumor size were observed in regular follow-up after one-month treatment. Results: All the patients were followed up for 6-14 months,and the postoperative recovery was good with no complication. One month after the implantation, the pain symptom was alleviated entirely in two nodes and partly in nine nodes. The tumor size shrank in ten nodes while there was no change in one node after one month. Conclusion: Permanent 125 I seeds interstitial implants for metastatic cervical cancer is a safe, minimally invasive and effective treatment. (authors)

  9. Accelerated partial-breast irradiation with interstitial implants. Analysis of factors affecting cosmetic outcome

    International Nuclear Information System (INIS)

    Ott, Oliver J.; Lotter, Michael; Fietkau, Rainer; Strnad, Vratislav

    2009-01-01

    Purpose: To analyze patient-, disease-, and treatment-related factors for their impact on cosmetic outcome (CO) after interstitial multicatheter accelerated partial-breast irradiation (APBI). Patients and Methods: Between April 2001 and January 2005, 171 patients with early breast cancer were recruited in Erlangen for this subanalysis of the German-Austrian APBI phase II-trial. 58% (99/171) of the patients received pulsed-dose-rate (PDR), and 42% (72/171) high-dose-rate (HDR) brachytherapy. Prescribed reference dose for HDR brachytherapy was 32 Gy in eight fractions of 4 Gy, twice daily. Prescribed reference dose in PDR brachytherapy was 49.8 Gy in 83 consecutive fractions of 0.6 Gy each hour. Total treatment time was 3-4 days. Endpoint of this evaluation was the CO, graded as excellent, good, fair, or poor. Patients were divided in two groups with an excellent (n = 102) or nonexcellent (n = 69) cosmetic result. Various factors were analyzed for their impact on excellent CO. Results: The median follow-up time was 52 months (range: 21-91 months). Cosmetic results were rated as excellent in 59.6% (102/171), good in 29.8% (51/171), fair in 9.9% (17/171), and poor in 0.6% (1/171). The initial cosmetic status was significantly worse for the nonexcellent CO group (p = 0.000). The percentage of patients who received PDR brachytherapy APBI was higher in the nonexcellent CO group (68.1% vs. 51%; p = 0.026). Acute toxicity was higher in the nonexcellent CO group (24.6% vs. 12.7%; p = 0.045). Furthermore, the presence of any late toxicity was found to be associated with a worse cosmetic result (65.2% vs. 18.6%; p = 0.000). In detail, the appearance of skin hyperpigmentation (p 0.034), breast tissue fibrosis (p = 0.000), and telangiectasia (p = 0.000) had a negative impact on CO. Conclusion: The initial, surgery-associated cosmetic status, brachytherapy modality, and the presence of acute and late toxicities were found to have an impact on overall CO. Our data have proven that

  10. A fast inverse treatment planning strategy facilitating optimized catheter selection in image-guided high-dose-rate interstitial gynecologic brachytherapy.

    Science.gov (United States)

    Guthier, Christian V; Damato, Antonio L; Hesser, Juergen W; Viswanathan, Akila N; Cormack, Robert A

    2017-12-01

    Interstitial high-dose rate (HDR) brachytherapy is an important therapeutic strategy for the treatment of locally advanced gynecologic (GYN) cancers. The outcome of this therapy is determined by the quality of dose distribution achieved. This paper focuses on a novel yet simple heuristic for catheter selection for GYN HDR brachytherapy and their comparison against state of the art optimization strategies. The proposed technique is intended to act as a decision-supporting tool to select a favorable needle configuration. The presented heuristic for catheter optimization is based on a shrinkage-type algorithm (SACO). It is compared against state of the art planning in a retrospective study of 20 patients who previously received image-guided interstitial HDR brachytherapy using a Syed Neblett template. From those plans, template orientation and position are estimated via a rigid registration of the template with the actual catheter trajectories. All potential straight trajectories intersecting the contoured clinical target volume (CTV) are considered for catheter optimization. Retrospectively generated plans and clinical plans are compared with respect to dosimetric performance and optimization time. All plans were generated with one single run of the optimizer lasting 0.6-97.4 s. Compared to manual optimization, SACO yields a statistically significant (P ≤ 0.05) improved target coverage while at the same time fulfilling all dosimetric constraints for organs at risk (OARs). Comparing inverse planning strategies, dosimetric evaluation for SACO and "hybrid inverse planning and optimization" (HIPO), as gold standard, shows no statistically significant difference (P > 0.05). However, SACO provides the potential to reduce the number of used catheters without compromising plan quality. The proposed heuristic for needle selection provides fast catheter selection with optimization times suited for intraoperative treatment planning. Compared to manual optimization, the

  11. Radiation Exposure Reduction to Brachytherapy Staff By Using Remote Afterloading

    International Nuclear Information System (INIS)

    Attalla, E.M.

    2005-01-01

    The radiation exposures to the personnel staff from patients with brachytherapy implants in a brachytherapy service were reviewed. Exposures to the brachytherapy personnel, as determined by Thermoluminescence Dosimeter (TLD) monitors, indicates a four-fold reduction in exposures after the implantation of the use of remote afterloading devices. Quarterly TLD monitor data for seven quarters prior to the use of remote afterloading devices demonstrate an average projected annual dose equivalent to the brachytherapy staff of 2543 Μ Sv. After the implantation of the remote afterloading devices, the quarterly TLD monitor data indicate an average dose equivalent per person of 153 Μ Sv. This is 76% reduction in exposure to brachytherapy personnel with the use of these devices

  12. Radium needles implant in the treatment of extensive vaginal involvement from cervical carcinoma

    International Nuclear Information System (INIS)

    Sewchand, W.; Prempree, T.; Patanaphan, V.; Carbone, D.; Salazar, O.M.

    1984-01-01

    An appraisal of the dosimetry of a modified brachytherapy approach is presented for improving the local control of extensive vaginal involvement from carcinoma of the cervix. This approach incorporates radium needles implant to the vaginal disease in conjunction with the usual routine intracavitary radium application. The aim of the interstitial implant is specifically to supplement the dose to the vaginal disease from the intracavitary application. Our procedure for accomplishing this boost in the dose to the vagina depends on the location, extent and thickness of the vaginal lesion following external beam irradiation of the whole pelvis. An increase of greater than 50% in the dose to the vaginal disease is gained by this combination intracavitary/implant approach which has been used in a variety of cases covering virtually all pertinent stages of cervical carcinoma. Discussion of the dosimetry of example cases is presented to demonstrate the value of combining interstitial and intracavitary therapy for this specific clinical application. (orig.)

  13. Prostate brachytherapy in Ghana: our initial experience

    Directory of Open Access Journals (Sweden)

    James Edward Mensah

    2016-10-01

    Full Text Available Purpose: This study presents the experience of a brachytherapy team in Ghana with a focus on technology transfer and outcome. The team was initially proctored by experienced physicians from Europe and South Africa. Material and methods : A total of 90 consecutive patients underwent either brachytherapy alone or brachytherapy in combination with external beam radiotherapy for prostate carcinoma between July 2008 and February 2014 at Korle Bu Teaching Hospital, Accra, Ghana. Patients were classified as low-risk, intermediate, and high-risk according to the National Comprehensive Cancer Network (NCCN criteria. All low-risk and some intermediate risk group patients were treated with seed implantation alone. Some intermediate and all high-risk group patients received brachytherapy combined with external beam radiotherapy. Results: The median patient age was 64.0 years (range 46-78 years. The median follow-up was 58 months (range 18-74 months. Twelve patients experienced biochemical failure including one patient who had evidence of metastatic disease and died of prostate cancer. Freedom from biochemical failure rates for low, intermediate, and high-risk cases were 95.4%, 90.9%, and 70.8%, respectively. Clinical parameters predictive of biochemical outcome included: clinical stage, Gleason score, and risk group. Pre-treatment prostate specific antigen (PSA was not a statistically significant predictor of biochemical failure. Sixty-nine patients (76.6% experienced grade 1 urinary symptoms in the form of frequency, urgency, and poor stream. These symptoms were mostly self-limiting. Four patients needed catheterization for urinary retention (grade 2. One patient developed a recto urethral fistula (grade 3 following banding for hemorrhoids. Conclusions : Our results compare favorably with those reported by other institutions with more extensive experience. We believe therefore that, interstitial permanent brachytherapy can be safely and effectively

  14. Implantation of xenon in amorphous carbon and silicon for brachytherapy application

    International Nuclear Information System (INIS)

    Marques, F.C.; Barbieri, P.F.; Viana, G.A.; Silva, D.S. da

    2013-01-01

    We report a procedure to implant high dose of xenon atoms (Xe) in amorphous carbon, a-C, and amorphous silicon, a-Si, for application in brachytherapy seeds. An ion beam assisted deposition (IBAD) system was used for the deposition of the films, where one ion gun was used for sputtering a carbon (or silicon) target, while the other ion gun was used to simultaneously bombard the growing film with a beam of xenon ion Xe + in the 0–300 eV range. Xe atoms were implanted into the film with concentration up to 5.5 at.%, obtained with Xe bombardment energy in the 50–150 eV range. X-ray absorption spectroscopy was used to investigate the local arrangement of the implanted Xe atoms through the Xe L III absorption edge (4.75 keV). It was observed that Xe atoms tend to agglomerate in nanoclusters in a-C and are dispersed in a-Si.

  15. Patterns of brachytherapy practice for patients with carcinoma of the cervix (1996-1999): A Patterns of Care Study

    International Nuclear Information System (INIS)

    Erickson, Beth; Eifel, Patricia; Moughan, Jennifer; Rownd, Jason M.S.; Iarocci, Thomas; Owen, Jean

    2005-01-01

    .5% and 9.4% of brachytherapy procedures included HDR or interstitial, respectively; in facilities that treated fewer patients, 3.4% had HDR brachytherapy, and only 1.2% had interstitial brachytherapy. Patients treated with LDR intracavitary radiotherapy had one (23.5%), two (74.1%), or three (2.4%) implants. For patients treated with curative intent who completed radiation therapy with LDR intracavitary radiation therapy without hysterectomy, the median brachytherapy dose to Point A was 40.3 Gy, and the median total dose to Point A was 82.9 Gy. Patients were treated with HDR intracavitary radiation therapy using a variety of treatment schedules using 1-2 fractions (7.5%), 3-4 fractions (17.4%), 5-6 fractions (38.5%), 7-9 fractions (33.5%), or 12 fractions (3%). Fraction sizes were 600 (8%), or unknown (9.2%). For patients treated with HDR, the median total dose to Point A (corrected for fraction size using a α/β = 10) was 85.8 Gy (range: 56.2-116.1 Gy). At institutions treating 56 days was significantly greater than at institutions ≥500 new patients per year (p = 0.002). Of the patients who had LDR intracavitary radiation therapy implants, 65% were treated using tandem and shielded Fletcher-Suit-Delclos colpostats; other patients had mini ovoids (10.9%), cylinders (3.9%), Henschke (3.7%), or other/mixed applicators (16.5%) . In contrast, of patients treated with HDR intracavitary radiation therapy, 68.7% had tandem and rings, 18.2% Fletcher-Suit-Delclos ovoids, 7.5% mini ovoids, 2.3% cylinders, and 3.2% other or mixed applicators. Conclusions: The median duration of treatment and median Point A dose were very similar for patients treated with HDR or LDR. Patients with HDR were treated using a variety of treatment schedules. Different applicator types were favored for LDR vs. HDR. Of patients treated with HDR, 73.4% had no brachytherapy bladder or rectal doses recorded, suggesting that full dosimetric calculations were performed only for the first fraction in many

  16. Use of Ir192 interstitial brachytherapy for an equine malignant dermal schwannoma : clinical communication

    Directory of Open Access Journals (Sweden)

    M.N. Saulez

    2009-05-01

    Full Text Available A 10-year-old Hanoverian mare was evaluated for a right buccal swelling that recurred 3 months following surgical resection. Ultrasonographic examination showed a broadly pedunculated subcutaneous mass at the level of 106-109 and 406-409 cheek teeth associated with an erosive mucosal lesion on the inside of the cheek. Histological examination of a biopsy specimen revealed a well-demarcated, malignant, dermal schwannoma. Following subcutaneous placement of platinum coated Ir192 wires under general anaesthesia, low-dose radiation of 5 gray per day was delivered for 14 days. Short-term complications included loss of patency of the right nasolacrimal duct, erythema, dermatitis, leukotrichia and left-sided deviation of the muzzle. Ten months later, there has been no tumour recurrence. Findings suggest that the use of interstitial brachytherapy should be considered for a malignant, dermal schwannoma that has recurred or is not amenable to surgery.

  17. {sup 125}I seed implant brachytherapy for the treatment of parotid gland cancers in children and adolescents

    Energy Technology Data Exchange (ETDEWEB)

    Zheng, L.; Zhang, J.; Song, T.; Zhang, J.; Yu, G.; Zhang, Y. [Peking University School and Hospital of Stomatology, Beijing (China). Dept. of Oral and Maxillofacial Surgery

    2013-05-15

    Background and purpose: There is a lack of optimal treatment strategies for managing salivary gland cancers in children and adolescents. This study is aimed at assessing the effect of {sup 125}I seed implantation for the treatment of parotid cancers in children and adolescents. Patients and methods: A total of 12 patients younger than 16 years with parotid gland malignant tumors underwent {sup 125}I seed implant brachytherapy between October 2003 and November 2008. All patients were assessed after treatment and at the local tumor control appointments. Facial nerve function, maxillofacial development, and radioactive side-effects were assessed. Results: The follow-up period ranged from 41-104 months. One patient with T4b died of pulmonary metastasis. The other patients were alive during the follow-up period. There were no serious radiation-related complications. The treatment did not affect facial nerve function and dentofacial growth in any of the children. Conclusion: For parotid gland cancers in children, {sup 125}I seed implant brachytherapy may be an acceptable treatment without serious complications and with satisfactory short-term effects. (orig.)

  18. Reduction of transient diffusion from 1 endash 5 keV Si+ ion implantation due to surface annihilation of interstitials

    International Nuclear Information System (INIS)

    Agarwal, A.; Gossmann, H.-.; Eaglesham, D.J.; Pelaz, L.; Jacobson, D.C.; Haynes, T.E.; Erokhin, Y.E.

    1997-01-01

    The reduction of transient enhanced diffusion (TED) with reduced implantation energy has been investigated and quantified. A fixed dose of 1x10 14 cm -2 Si + was implanted at energies ranging from 0.5 to 20 keV into boron doping superlattices and enhanced diffusion of the buried boron marker layers was measured for anneals at 810, 950, and 1050 degree C. A linearly decreasing dependence of diffusivity enhancement on decreasing Si + ion range is observed at all temperatures, extrapolating to ∼1 for 0 keV. This is consistent with our expectation that at zero implantation energy there would be no excess interstitials from the implantation and hence no TED. Monte Carlo modeling and continuum simulations are used to fit the experimental data. The results are consistent with a surface recombination length for interstitials of <10 nm. The data presented here demonstrate that in the range of annealing temperatures of interest for p-n junction formation, TED is reduced at smaller ion implantation energies and that this is due to increased interstitial annihilation at the surface. copyright 1997 American Institute of Physics

  19. MO-E-BRD-02: Accelerated Partial Breast Irradiation in Brachytherapy: Is Shorter Better?

    International Nuclear Information System (INIS)

    Todor, D.

    2015-01-01

    Is Non-invasive Image-Guided Breast Brachytherapy Good? – Jess Hiatt, MS Non-invasive Image-Guided Breast Brachytherapy (NIBB) is an emerging therapy for breast boost treatments as well as Accelerated Partial Breast Irradiation (APBI) using HDR surface breast brachytherapy. NIBB allows for smaller treatment volumes while maintaining optimal target coverage. Considering the real-time image-guidance and immobilization provided by the NIBB modality, minimal margins around the target tissue are necessary. Accelerated Partial Breast Irradiation in brachytherapy: is shorter better? - Dorin Todor, PhD VCU A review of balloon and strut devices will be provided together with the origins of APBI: the interstitial multi-catheter implant. A dosimetric and radiobiological perspective will help point out the evolution in breast brachytherapy, both in terms of devices and the protocols/clinical trials under which these devices are used. Improvements in imaging, delivery modalities and convenience are among the factors driving the ultrashort fractionation schedules but our understanding of both local control and toxicities associated with various treatments is lagging. A comparison between various schedules, from a radiobiological perspective, will be given together with a critical analysis of the issues. to review and understand the evolution and development of APBI using brachytherapy methods to understand the basis and limitations of radio-biological ‘equivalence’ between fractionation schedules to review commonly used and proposed fractionation schedules Intra-operative breast brachytherapy: Is one stop shopping best?- Bruce Libby, PhD. University of Virginia A review of intraoperative breast brachytherapy will be presented, including the Targit-A and other trials that have used electronic brachytherapy. More modern approaches, in which the lumpectomy procedure is integrated into an APBI workflow, will also be discussed. Learning Objectives: To review past and current

  20. MO-E-BRD-02: Accelerated Partial Breast Irradiation in Brachytherapy: Is Shorter Better?

    Energy Technology Data Exchange (ETDEWEB)

    Todor, D. [Virginia Commonwealth University (United States)

    2015-06-15

    Is Non-invasive Image-Guided Breast Brachytherapy Good? – Jess Hiatt, MS Non-invasive Image-Guided Breast Brachytherapy (NIBB) is an emerging therapy for breast boost treatments as well as Accelerated Partial Breast Irradiation (APBI) using HDR surface breast brachytherapy. NIBB allows for smaller treatment volumes while maintaining optimal target coverage. Considering the real-time image-guidance and immobilization provided by the NIBB modality, minimal margins around the target tissue are necessary. Accelerated Partial Breast Irradiation in brachytherapy: is shorter better? - Dorin Todor, PhD VCU A review of balloon and strut devices will be provided together with the origins of APBI: the interstitial multi-catheter implant. A dosimetric and radiobiological perspective will help point out the evolution in breast brachytherapy, both in terms of devices and the protocols/clinical trials under which these devices are used. Improvements in imaging, delivery modalities and convenience are among the factors driving the ultrashort fractionation schedules but our understanding of both local control and toxicities associated with various treatments is lagging. A comparison between various schedules, from a radiobiological perspective, will be given together with a critical analysis of the issues. to review and understand the evolution and development of APBI using brachytherapy methods to understand the basis and limitations of radio-biological ‘equivalence’ between fractionation schedules to review commonly used and proposed fractionation schedules Intra-operative breast brachytherapy: Is one stop shopping best?- Bruce Libby, PhD. University of Virginia A review of intraoperative breast brachytherapy will be presented, including the Targit-A and other trials that have used electronic brachytherapy. More modern approaches, in which the lumpectomy procedure is integrated into an APBI workflow, will also be discussed. Learning Objectives: To review past and current

  1. MO-E-BRD-03: Intra-Operative Breast Brachytherapy: Is One Stop Shopping Best?

    International Nuclear Information System (INIS)

    Libby, B.

    2015-01-01

    Is Non-invasive Image-Guided Breast Brachytherapy Good? – Jess Hiatt, MS Non-invasive Image-Guided Breast Brachytherapy (NIBB) is an emerging therapy for breast boost treatments as well as Accelerated Partial Breast Irradiation (APBI) using HDR surface breast brachytherapy. NIBB allows for smaller treatment volumes while maintaining optimal target coverage. Considering the real-time image-guidance and immobilization provided by the NIBB modality, minimal margins around the target tissue are necessary. Accelerated Partial Breast Irradiation in brachytherapy: is shorter better? - Dorin Todor, PhD VCU A review of balloon and strut devices will be provided together with the origins of APBI: the interstitial multi-catheter implant. A dosimetric and radiobiological perspective will help point out the evolution in breast brachytherapy, both in terms of devices and the protocols/clinical trials under which these devices are used. Improvements in imaging, delivery modalities and convenience are among the factors driving the ultrashort fractionation schedules but our understanding of both local control and toxicities associated with various treatments is lagging. A comparison between various schedules, from a radiobiological perspective, will be given together with a critical analysis of the issues. to review and understand the evolution and development of APBI using brachytherapy methods to understand the basis and limitations of radio-biological ‘equivalence’ between fractionation schedules to review commonly used and proposed fractionation schedules Intra-operative breast brachytherapy: Is one stop shopping best?- Bruce Libby, PhD. University of Virginia A review of intraoperative breast brachytherapy will be presented, including the Targit-A and other trials that have used electronic brachytherapy. More modern approaches, in which the lumpectomy procedure is integrated into an APBI workflow, will also be discussed. Learning Objectives: To review past and current

  2. BEDVH--A method for evaluating biologically effective dose volume histograms: Application to eye plaque brachytherapy implants

    International Nuclear Information System (INIS)

    Gagne, Nolan L.; Leonard, Kara L.; Huber, Kathryn E.; Mignano, John E.; Duker, Jay S.; Laver, Nora V.; Rivard, Mark J.

    2012-01-01

    Purpose: A method is introduced to examine the influence of implant duration T, radionuclide, and radiobiological parameters on the biologically effective dose (BED) throughout the entire volume of regions of interest for episcleral brachytherapy using available radionuclides. This method is employed to evaluate a particular eye plaque brachytherapy implant in a radiobiological context. Methods: A reference eye geometry and 16 mm COMS eye plaque loaded with 103 Pd, 125 I, or 131 Cs sources were examined with dose distributions accounting for plaque heterogeneities. For a standardized 7 day implant, doses to 90% of the tumor volume ( TUMOR D 90 ) and 10% of the organ at risk volumes ( OAR D 10 ) were calculated. The BED equation from Dale and Jones and published α/β and μ parameters were incorporated with dose volume histograms (DVHs) for various T values such as T = 7 days (i.e., TUMOR 7 BED 10 and OAR 7 BED 10 ). By calculating BED throughout the volumes, biologically effective dose volume histograms (BEDVHs) were developed for tumor and OARs. Influence of T, radionuclide choice, and radiobiological parameters on TUMOR BEDVH and OAR BEDVH were examined. The nominal dose was scaled for shorter implants to achieve biological equivalence. Results: TUMOR D 90 values were 102, 112, and 110 Gy for 103 Pd, 125 I, and 131 Cs, respectively. Corresponding TUMOR 7 BED 10 values were 124, 140, and 138 Gy, respectively. As T decreased from 7 to 0.01 days, the isobiologically effective prescription dose decreased by a factor of three. As expected, TUMOR 7 BEDVH did not significantly change as a function of radionuclide half-life but varied by 10% due to radionuclide dose distribution. Variations in reported radiobiological parameters caused TUMOR 7 BED 10 to deviate by up to 46%. Over the range of OAR α/β values, OAR 7 BED 10 varied by up to 41%, 3.1%, and 1.4% for the lens, optic nerve, and lacrimal gland, respectively. Conclusions: BEDVH permits evaluation of the

  3. Sexual function after permanent prostate brachytherapy

    International Nuclear Information System (INIS)

    Galbreath, R.W.; Merrick, G.S.; Butler, W.M.; Stipetich, R.L.; Abel, L.J.; Lief, J.H.

    2001-01-01

    Purpose: To determine the incidence of potency preservation following permanent prostate brachytherapy and to evaluate the effect of multiple clinical and treatment parameters on penile erectile function. Materials and Methods: 425 patients underwent permanent prostate brachytherapy from April 1995 to October 1999. 209 patients who were potent prior to brachytherapy and currently not receiving hormonal manipulation were mailed an International Index of Erectile Function (IIEF) questionnaire with a pre-addressed stamped envelope. 180 patients completed and returned the questionnaire. Median patient follow-up was 39 months (range 18-74 months). Pre-implant erectile function was assigned using a three-tiered scoring system (2 = erections always or nearly always sufficient for vaginal penetration; 1 = erections sufficient for vaginal penetration but considered suboptimal; 0 = the inability to obtain erections and/or erections inadequate for vaginal penetration). Post-implant potency was defined as an IIEF score >11. Clinical parameters evaluated for sexual function included patient age, clinical T stage, elapsed time since implantation, hypertension, diabetes mellitus, and tobacco consumption. Evaluated treatment parameters included the utilization of neoadjuvant hormonal manipulation and the choice of isotope. The efficacy of sildenafil citrate in brachytherapy induced erectile dysfunction (ED) was also evaluated. Results: A pre-treatment erectile function score of 2 and 1 were assigned to 126 and 54 patients respectively. With 6 year follow up, 39% of patients maintained potency following prostate brachytherapy with a plateau on the curve. Post-implant preservation of potency (IIEF>11) correlated with pre-implant erectile function (50% versus 14% for pre-implant scores of 2 and 1 respectively, p≤0.0001), patient age (56%, 38%, and 23% for patients <60 years of age, 60-69 years of age, and ≥70 years of age respectively, p=0.012) and a history of diabetes mellitus

  4. Feasibility and early results of interstitial intensity-modulated HDR/PDR brachytherapy (IMBT) with/without complementary external-beam radiotherapy and extended surgery in recurrent pelvic colorectal cancer

    International Nuclear Information System (INIS)

    Tepel, J.; Bokelmann, F.; Faendrich, F.; Kremer, B.; Schmid, A.; Kovacs, G.; University Hospital of Schleswig-Holstein, Kiel

    2005-01-01

    Background: A new multimodality treatment concept consisting of extended resection and postoperative fractionated intensity-modulated interstitial brachytherapy (IMBT) was introduced for pelvic recurrence of colorectal carcinoma. Patients and Methods: 46 patients received extended resection and single plastic tubes were sutured directly onto the tumor bed. IMBT was started within 2 weeks postoperatively with a median dose of 24.5 Gy (5-35 Gy). Patients were treated either with high-dose-rate brachytherapy (HDR; n=23) or with pulsed-dose-rate brachytherapy (PDR; n=23). 25 patients received complementary 45-Gy external-beam irradiation (EBRT) to the pelvic region after explanting the plastic tubes. Results: Median follow-up was 20.6 months (7-107 months) and mean patient survival 25.7±25.8 months (median 17, range 1-107 months). After 5 years overall survival, disease-free survival and local control rate were 23%, 20% and 33%, significantly influenced by the resectional state. There was a trend in favor of PDR compared to HDR, which reached statistical significance in patients who had not received additional EBRT. Conclusion: The combination of extended surgery and postoperative interstitial IMBT is feasible and offers effective interdisciplinary treatment of recurrent colorectal cancer. In this small and inhomogeneous cohort of patients PDR seems to be more effective than HDR, particularly when application of complementary EBRT is not possible. None of the patients who required resection of distant metastasis survived >2 years in this study. (orig.)

  5. WE-AB-BRA-11: Improved Imaging of Permanent Prostate Brachytherapy Seed Implants by Combining an Endorectal X-Ray Sensor with a CT Scanner

    International Nuclear Information System (INIS)

    Steiner, J; Matthews, K; Jia, G

    2016-01-01

    Purpose: To test feasibility of the use of a digital endorectal x-ray sensor for improved image resolution of permanent brachytherapy seed implants compared to conventional CT. Methods: Two phantoms simulating the male pelvic region were used to test the capabilities of a digital endorectal x-ray sensor for imaging permanent brachytherapy seed implants. Phantom 1 was constructed from acrylic plastic with cavities milled in the locations of the prostate and the rectum. The prostate cavity was filled a Styrofoam plug implanted with 10 training seeds. Phantom 2 was constructed from tissue-equivalent gelatins and contained a prostate phantom implanted with 18 strands of training seeds. For both phantoms, an intraoral digital dental x-ray sensor was placed in the rectum within 2 cm of the seed implants. Scout scans were taken of the phantoms over a limited arc angle using a CT scanner (80 kV, 120–200 mA). The dental sensor was removed from the phantoms and normal helical CT and scout (0 degree) scans using typical parameters for pelvic CT (120 kV, auto-mA) were collected. A shift-and add tomosynthesis algorithm was developed to localize seed plane location normal to detector face. Results: The endorectal sensor produced images with improved resolution compared to CT scans. Seed clusters and individual seed geometry were more discernable using the endorectal sensor. Seed 3D locations, including seeds that were not located in every projection image, were discernable using the shift and add algorithm. Conclusion: This work shows that digital endorectal x-ray sensors are a feasible method for improving imaging of permanent brachytherapy seed implants. Future work will consist of optimizing the tomosynthesis technique to produce higher resolution, lower dose images of 1) permanent brachytherapy seed implants for post-implant dosimetry and 2) fine anatomic details for imaging and managing prostatic disease compared to CT images. Funding: LSU Faculty Start-up Funding

  6. WE-AB-BRA-11: Improved Imaging of Permanent Prostate Brachytherapy Seed Implants by Combining an Endorectal X-Ray Sensor with a CT Scanner

    Energy Technology Data Exchange (ETDEWEB)

    Steiner, J; Matthews, K; Jia, G [Louisiana State University, Baton Rouge, LA (United States)

    2016-06-15

    Purpose: To test feasibility of the use of a digital endorectal x-ray sensor for improved image resolution of permanent brachytherapy seed implants compared to conventional CT. Methods: Two phantoms simulating the male pelvic region were used to test the capabilities of a digital endorectal x-ray sensor for imaging permanent brachytherapy seed implants. Phantom 1 was constructed from acrylic plastic with cavities milled in the locations of the prostate and the rectum. The prostate cavity was filled a Styrofoam plug implanted with 10 training seeds. Phantom 2 was constructed from tissue-equivalent gelatins and contained a prostate phantom implanted with 18 strands of training seeds. For both phantoms, an intraoral digital dental x-ray sensor was placed in the rectum within 2 cm of the seed implants. Scout scans were taken of the phantoms over a limited arc angle using a CT scanner (80 kV, 120–200 mA). The dental sensor was removed from the phantoms and normal helical CT and scout (0 degree) scans using typical parameters for pelvic CT (120 kV, auto-mA) were collected. A shift-and add tomosynthesis algorithm was developed to localize seed plane location normal to detector face. Results: The endorectal sensor produced images with improved resolution compared to CT scans. Seed clusters and individual seed geometry were more discernable using the endorectal sensor. Seed 3D locations, including seeds that were not located in every projection image, were discernable using the shift and add algorithm. Conclusion: This work shows that digital endorectal x-ray sensors are a feasible method for improving imaging of permanent brachytherapy seed implants. Future work will consist of optimizing the tomosynthesis technique to produce higher resolution, lower dose images of 1) permanent brachytherapy seed implants for post-implant dosimetry and 2) fine anatomic details for imaging and managing prostatic disease compared to CT images. Funding: LSU Faculty Start-up Funding

  7. The Syed temporary interstitial iridium gynaecological implant: an inverse planning system

    International Nuclear Information System (INIS)

    Fung, Albert Y.C.

    2002-01-01

    Patients with advanced gynaecological cancer are often treated with a temporary interstitial implant using the Syed template and Ir-192 ribbons at the Memorial Sloan-Kettering Cancer Center. Urgency in planning is great. We created a computerized inverse planning system for the Syed temporary gynaecological implant, which optimized the ribbon strengths a few seconds after catheter digitization. Inverse planning was achieved with simulated annealing. We discovered that hand-drawn target volumes had drawbacks; hence instead of producing a grid of points based on target volume, the optimization points were generated directly from the catheter positions without requiring an explicit target volume. Since all seeds in the same ribbon had the same strength, the minimum doses were located at both ends of the implant. Optimization points generated at both ends ensured coverage of the whole implant. Inverse planning took only a few seconds, and generated plans that provide a good starting point for manual improvement. (author)

  8. Is there any advantage of CT based 3-dimensional conformal planning over conventional orthogonal x-ray based planning in HDR brachytherapy in breast cancer

    International Nuclear Information System (INIS)

    Biswal, B.M.; Idris, N.R.; Zakaria, A.B.; Khairul, N.

    2003-01-01

    The conventional brachytherapy dose calculation is based on a particular brachytherapy rule or individual dosimetry based on the reconstruction of the sources from the orthogonal films. In the recent years many centers are using CT based 3D conformal brachytherapy in order to improve the dosimetric outcome of a given plan. Here we would like to present our experience on the use of both techniques to deliver HDR interstitial brachytherapy as boost in early breast cancer. From January 2001 to January 2003, we treated 4 breast cancer patients using conventional orthogonal x-rays and CT scan in 3 cases for the treatment plan. All patients received an external beam radiotherapy dose of 46 Gy in 23 fractions over 4.5 weeks to the whole breast using 6 MV photon beam. Subsequently the primary lesion was supplimented with HDR brachytherapy to a dose of 2.5 Gy BID for 3 consecutive days using a (192)Ir microSelectronHDR. The dose prescription was individualized to encompass the tumor volume with a 10 mm margin. The differences of the dosimetric outcome were compared. All patients completed above schedule of radiotherapy. The primary was implanted with single plane in 3 patients and multiplane implant in 4 patients. Orthogonal x-ray based localization was performed in 4 patients and CT scan based localization in 3 cases. Three patients were implanted single plane and 4 patients with multiplane implants with a median catheter number of 9 (range 6-14). The 3D conformal dose optimization was performed using Nucletron planning system (Plato). The mean 100% and 150% isodose volume was 67.3 cm 3 and 31.25cm 3 respectively. The identification of primary tumor volume, organ at risk, and identification of afterloading catheters were superior in CT based plan than conventional planning. CT scan based 3D conformal brachytherapy planning give better identification of tumor volume and its curvature, decrease the time to identify the sources and evaluate the radiation dose to organs at

  9. Role of brachytherapy in the treatment of localized prostate cancer

    Directory of Open Access Journals (Sweden)

    A. D. Kaprin

    2015-01-01

    Full Text Available The review is devoted to application of brachytherapy for treating the localized prostate cancer (PC. Statistics for incidence and detectability of this pathology and its dynamics for recent years are represented. Brief analysis of other methods which are conveniently used for treatment of PC, such as radical prostatectomy and external-beam radiotherapy, was performed. Advantages and disadvantages of these methods have been discussed. Brief history about the development of brachytherapy from first experience to wide-spread use in clinical practice is reported. The detailed review of series of large trials from Russia and other countries for efficiency and safety of brachytherapy in patients with prostate cancer for recent 15 years is also represented. Two types of brachytherapy in current clinical oncology i.e. low-dose technique with permanent implantation of microsources and high-dose temporary isotope implantation, specifics of its application in different groups of patients have been described. The procedure of brachytherapy and its three main steps i.e. planning, implantation and control assessment after implantation have been characterized in details. The conclusion about benefits of using of brachytherapy in the treatment of prostate cancer as minimally invasive and efficient method was made. 

  10. Automated treatment planning engine for prostate seed implant brachytherapy

    International Nuclear Information System (INIS)

    Yu Yan; Zhang, J.B.Y.; Brasacchio, Ralph A.; Okunieff, Paul G.; Rubens, Deborah J.; Strang, John G.; Soni, Arvind; Messing, Edward M.

    1999-01-01

    Purpose: To develop a computer-intelligent planning engine for automated treatment planning and optimization of ultrasound- and template-guided prostate seed implants. Methods and Materials: The genetic algorithm was modified to reflect the 2D nature of the implantation template. A multi-objective decision scheme was used to rank competing solutions, taking into account dose uniformity and conformity to the planning target volume (PTV), dose-sparing of the urethra and the rectum, and the sensitivity of the resulting dosimetry to seed misplacement. Optimized treatment plans were evaluated using selected dosimetric quantifiers, dose-volume histogram (DVH), and sensitivity analysis based on simulated seed placement errors. These dosimetric planning components were integrated into the Prostate Implant Planning Engine for Radiotherapy (PIPER). Results: PIPER has been used to produce a variety of plans for prostate seed implants. In general, maximization of the minimum peripheral dose (mPD) for given implanted total source strength tended to produce peripherally weighted seed patterns. Minimization of the urethral dose further reduced the loading in the central region of the PTV. Isodose conformity to the PTV was achieved when the set of objectives did not reflect seed positioning uncertainties; the corresponding optimal plan generally required fewer seeds and higher source strength per seed compared to the manual planning experience. When seed placement uncertainties were introduced into the set of treatment planning objectives, the optimal plan tended to reach a compromise between the preplanned outcome and the likelihood of retaining the preferred outcome after implantation. The reduction in the volatility of such seed configurations optimized under uncertainty was verified by sensitivity studies. Conclusion: An automated treatment planning engine incorporating real-time sensitivity analysis was found to be a useful tool in dosimetric planning for prostate

  11. Effect of permanent 103Pd radioactive seed implantation on brachytherapy of malignant tumor

    International Nuclear Information System (INIS)

    Chen Ping; Wei Xianzhong; Liu Yanmin; Wu Kaijun; Liang Jianxin; Chen Hanzhang

    2003-01-01

    Objective: To investigate and assess the brachytherapeutic effectiveness of 103 Pd radioactive seeds in malignant tumor therapy. Methods: 196.1-2127.5 MBq 103 Pd seeds were implanted in 21 confirmed malignant tumor patients. The seeds were evenly scattered in 15/21 patients' tumors and peripherally in the remaining 6 cases' tumors. The size and shape, local recurrence and remote metastasis of the tumors were observed. Results: The brachytherapy of 103 Pd seeds in tumor patients resulted in obvious efficacy. No local recurrence and remote metastasis were observed. 19/21 (90.5%) patients scored 0 and 2/21 (9.5%) of them scored 1 in skin acute radiation morbidity scoring criteria within the observation period. Conclusion: The 103 Pd seeds can be safely used in brachytherapy of malignant tumors with lower or medium sensitivity to radiation therapy

  12. Effect of post-implant edema on prostate brachytherapy treatment margins

    International Nuclear Information System (INIS)

    Reed, Daniel R.; Wallner, Kent; Ford, Eric; Mueller, Amy; Merrick, Gregory; Maki, Jeffrey; Sutlief, Steven; Butler, Wayne

    2005-01-01

    Purpose: To determine if postimplant prostate brachytherapy treatment margins calculated on Day 0 differ substantially from those calculated on Day 30. Methods: Thirty patients with 1997 American Joint Commission on Cancer clinical stage T1-T2 prostatic carcinoma underwent prostate brachytherapy with I-125 prescribed to 144 Gy. Treatment planning methods included using loose seeds in a modified peripheral loading pattern and treatment margins (TMs) of 5-8 mm. Postimplant plain radiographs, computed tomography scans, and magnetic resonance scans were obtained 1-4 hours after implantation (Day 0). A second set of imaging studies was obtained at 30 days after implantation (Day 30) and similarly analyzed. Treatment margins were measured as the radial distance in millimeters from the prostate edge to the 100% isodose line. The TMs were measured and tabulated at 90 o intervals around the prostate periphery at 0.6-cm intervals. Each direction was averaged to obtain the mean anterior, posterior, left, and right margins. Results: The mean overall TM increased from 2.6 mm (±2.3) on Day 0 to 3.5 mm (±2.4) on Day 30. The mean anterior margin increased from 1.2 mm on Day 0 to 1.8 mm on Day 30. The posterior margin increased from 1.2 mm on Day 0 to 2.8 mm on Day 30. The lateral treatment margins increased most over time, with mean right treatment margin increasing from 3.9 mm on Day 0 to 4.7 mm on Day 30. Conclusion: Treatment margins appear to be durable in the postimplant period, with a clinically insignificant increase from Day 0 to Day 30

  13. Three-dimensional viewing and dosimetric calculations of Au-198 implants of the prostate

    International Nuclear Information System (INIS)

    Avizonis, V.N.; Anderson, K.M.; Jani, S.K.; Hussey, D.H.

    1991-01-01

    Dose gradients for brachytherapy vary considerably in three dimensions, which complicates conventional two-dimensional dosimetry. Recent developments in computer graphics technology have enabled visualization of anatomy and radiation doses in three dimensions. The objective of this paper is to develop a three-dimensional viewing and dosimetry program for brachytherapy and to test this system in phantoms and in patients undergoing Au-198 implants in the prostate. Three-dimensional computer algorithms for the author's Silicon Graphics supercomputing workstation were developed, tested, and modified on the basis of studies in phantoms and patients. Studies were performed on phantoms of known dimensions and gold seeds in known locations to assess the accuracy of volume reconstruction, seed placement, and isodose distribution. Isodose curves generated with the three-dimensional system were compared with those generated by a Theratronics Treatment Planning Computer using conventional methods. Twenty patients with permanent Au-198 interstitial implants in the prostate were similarly studied

  14. Interstitial high-dose rate brachytherapy as boost for anal canal cancer

    International Nuclear Information System (INIS)

    Falk, Alexander Tuan; Claren, Audrey; Benezery, Karen; François, Eric; Gautier, Mathieu; Gerard, Jean-Pierre; Hannoun-Levi, Jean-Michel

    2014-01-01

    To assess clinical outcomes of patients treated with a high-dose rate brachytherapy boost for anal canal cancer (ACC). From August 2005 to February 2013, 28 patients presenting an ACC treated by split-course external beam radiotherapy (EBRT) and HDR brachytherapy with or without chemotherapy in a French regional cancer center in Nice were retrospectively analyzed. Median age was 60.6 years [34 – 83], 25 patients presented a squamous cell carcinoma and 3 an adenocarcinoma; 21 received chemotherapy. Median dose of EBRT was 45 Gy [43.2 – 52]. Median dose of HDR brachytherapy was 12 Gy [10 - 15] with a median duration of 2 days. Median overall treatment time was 63 days and median delay between EBRT and brachytherapy was 20 days. Two-year local relapse free, metastatic free, disease free and overall survivals were 83%, 81.9%, 71.8% and 87.7% respectively. Acute toxicities were frequent but not severe with mostly grade 1 toxicities: 37% of genito-urinary, 40.7% of gastro-intestinal and 3.7% of cutaneous toxicities. Late toxicities were mainly G1 (43.1%) and G2 (22%). Two-year colostomy-free survival was 75.1%, one patient had a definitive sphincter amputation. High-dose rate brachytherapy for anal canal carcinoma as boost represents a feasible technique compared to low or pulsed-dose rate brachytherapy. This technique remains an excellent approach to precisely boost the tumor in reducing the overall treatment time

  15. Primary calibration of coiled 103Pd brachytherapy sources

    International Nuclear Information System (INIS)

    Paxton, Adam B.; Culberson, Wesley S.; DeWerd, Larry A.; Micka, John A.

    2008-01-01

    Coiled 103 Pd brachytherapy sources have been developed by RadioMed Corporation for use as low-dose-rate (LDR) interstitial implants. The coiled sources are provided in integer lengths from 1 to 6 cm and address many common issues seen with traditional LDR brachytherapy sources. The current standard for determining the air-kerma strength (S K ) of low-energy LDR brachytherapy sources is the National Institute of Standards and Technology's Wide-Angle Free-Air Chamber (NIST WAFAC). Due to geometric limitations, however, the NIST WAFAC is unable to determine the S K of sources longer than 1 cm. This project utilized the University of Wisconsin's Variable-Aperture Free-Air Chamber (UW VAFAC) to determine the S K of the longer coiled sources. The UW VAFAC has shown agreement in S K values of 1 cm length coils to within 1% of those determined with the NIST WAFAC, but the UW VAFAC does not share the same geometric limitations as the NIST WAFAC. A new source holder was constructed to hold the coiled sources in place during measurements with the UW VAFAC. Correction factors for the increased length of the sources have been determined and applied to the measurements. Using the new source holder and corrections, the S K of 3 and 6 cm coiled sources has been determined. Corrected UW VAFAC data and ionization current measurements from well chambers have been used to determine calibration coefficients for use in the measurement of 3 and 6 cm coiled sources in well chambers. Thus, the UW VAFAC has provided the first transferable, primary measurement of low-energy LDR brachytherapy sources with lengths greater than 1 cm

  16. Prostate brachytherapy - discharge

    Science.gov (United States)

    Implant therapy - prostate cancer - discharge; Radioactive seed placement - discharge ... You had a procedure called brachytherapy to treat prostate cancer. Your treatment lasted 30 minutes or more, ...

  17. Image fusion techniques in permanent seed implantation

    Directory of Open Access Journals (Sweden)

    Alfredo Polo

    2010-10-01

    Full Text Available Over the last twenty years major software and hardware developments in brachytherapy treatment planning, intraoperative navigation and dose delivery have been made. Image-guided brachytherapy has emerged as the ultimate conformal radiation therapy, allowing precise dose deposition on small volumes under direct image visualization. In thisprocess imaging plays a central role and novel imaging techniques are being developed (PET, MRI-MRS and power Doppler US imaging are among them, creating a new paradigm (dose-guided brachytherapy, where imaging is used to map the exact coordinates of the tumour cells, and to guide applicator insertion to the correct position. Each of these modalities has limitations providing all of the physical and geometric information required for the brachytherapy workflow.Therefore, image fusion can be used as a solution in order to take full advantage of the information from each modality in treatment planning, intraoperative navigation, dose delivery, verification and follow-up of interstitial irradiation.Image fusion, understood as the visualization of any morphological volume (i.e. US, CT, MRI together with an additional second morpholo gical volume (i.e. CT, MRI or functional dataset (functional MRI, SPECT, PET, is a well known method for treatment planning, verification and follow-up of interstitial irradiation. The term image fusion is used when multiple patient image datasets are registered and overlaid or merged to provide additional information. Fused images may be created from multiple images from the same imaging modality taken at different moments (multi-temporalapproach, or by combining information from multiple modalities. Quality means that the fused images should provide additional information to the brachythe rapy process (diagnosis and staging, treatment planning, intraoperative imaging, treatment delivery and follow-up that cannot be obtained in other ways. In this review I will focus on the role of

  18. Penile brachytherapy: Results for 49 patients

    International Nuclear Information System (INIS)

    Crook, Juanita M.; Jezioranski, John; Grimard, Laval; Esche, Bernd; Pond, G.

    2005-01-01

    Purpose: To report results for 49 men with squamous cell carcinoma (SCC) of the penis treated with primary penile interstitial brachytherapy at one of two institutions: the Ottawa Regional Cancer Center, Ottawa, and the Princess Margaret Hospital, Toronto, Ontario, Canada. Methods and Materials: From September 1989 to September 2003, 49 men (mean age, 58 years; range, 22-93 years) had brachytherapy for penile SCC. Fifty-one percent of tumors were T1, 33% T2, and 8% T3; 4% were in situ and 4% Tx. Grade was well differentiated in 31%, moderate in 45%, and poor in 2%; grade was unspecified for 20%. One tumor was verrucous. All tumors in Toronto had pulsed dose rate (PDR) brachytherapy (n = 23), whereas those in Ottawa had either Iridium wire (n 22) or seeds (n = 4). Four patients had a single plane implant with a plastic tube technique, and all others had a volume implant with predrilled acrylic templates and two or three parallel planes of needles (median, six needles). Mean needle spacing was 13.5 mm (range, 10-18 mm), mean dose rate was 65 cGy/h (range, 33-160 cGy/h), and mean duration was 98.8 h (range, 36-188 h). Dose rates for PDR brachytherapy were 50-61.2 cGy/h, with no correction in total dose, which was 60 Gy in all cases. Results: Median follow-up was 33.4 months (range, 4-140 months). At 5 years, actuarial overall survival was 78.3% and cause-specific survival 90.0%. Four men died of penile cancer, and 6 died of other causes with no evidence of recurrence. The cumulative incidence rate for never having experienced any type of failure at 5 years was 64.4% and for local failure was 85.3%. All 5 patients with local failure were successfully salvaged by surgery; 2 other men required penectomy for necrosis. The soft tissue necrosis rate was 16% and the urethral stenosis rate 12%. Of 8 men with regional failure, 5 were salvaged by lymph node dissection with or without external radiation. All 4 men with distant failure died of disease. Of 49 men, 42 had an intact

  19. Radioablation of adrenal gland malignomas with interstitial high-dose-rate brachytherapy. Efficacy and outcome

    Energy Technology Data Exchange (ETDEWEB)

    Mohnike, K. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Radiologie und Nuklearmedizin, Magdeburg (Germany); DTZ am Frankfurter Tor, Berlin (Germany); Neumann, K.; Seidensticker, M.; Seidensticker, R.; Pech, M.; Streitparth, T.; Ricke, J. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Radiologie und Nuklearmedizin, Magdeburg (Germany); Hass, P.; Gademann, G. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Strahlentherapie, Magdeburg (Germany); Klose, S. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Nieren- und Hochdruckkrankheiten, Diabetologie und Endokrinologie, Magdeburg (Germany); Garlipp, B.; Benckert, C. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Allgemein-, Viszeral- und Gefaesschirurgie, Magdeburg (Germany); Wendler, J.J.; Liehr, U.B.; Schostak, M. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Urologie und Kinderurologie, Magdeburg (Germany); Goeppner, D. [Universitaetsklinikum Magdeburg A.oe.R., Otto-von-Guericke-Universitaet, Klinik fuer Dermatologie, Magdeburg (Germany)

    2017-08-15

    To assess the efficacy, safety, and outcome of image-guided high-dose-rate (HDR) brachytherapy in patients with adrenal gland metastases (AGM). From January 2007 to April 2014, 37 patients (7 female, 30 male; mean age 66.8 years, range 41.5-82.5 years) with AGM from different primary tumors were treated with CT-guided HDR interstitial brachytherapy (iBT). Primary endpoint was local tumor control (LTC). Secondary endpoints were time to untreatable progression (TTUP), time to progression (TTP), overall survival (OS), and safety. In a secondary analysis, risk factors with an influence on survival were identified. The median biological equivalent dose (BED) was 37.4 Gy. Mean LTC after 12 months was 88%; after 24 months this was 74%. According to CTCAE criteria, one grade 3 adverse event occurred. Median OS after first diagnosis of AGM was 18.3 months. Median OS, TTUP, and TTP after iBT treatment were 11.4, 6.6, and 3.5 months, respectively. Uni- and multivariate Cox regression analyses revealed significant influences of synchronous disease, tumor diameter, and the total number of lesions on OS or TTUP or both. Image-guided HDR-iBT is safe and effective. Treatment- and primary tumor-independent features influenced survival of patients with AGM after HDR-iBR treatment. (orig.) [German] Beurteilung der Effektivitaet, Sicherheit und Ergebnisse nach bildgefuehrter High-dose-rate-(HDR-)Brachytherapie bei Patienten mit Nebennierenmetastasen. Von Januar 2007 bis April 2014 wurden 37 Patienten (7 weiblich, 30 maennlich; mittleres Alter 66,8 Jahre, Spanne 41,5-82,5 Jahre) mit Nebennierenmetastasen verschiedener Primarien mit CT-gesteuerter interstitieller HDR-Brachytherapie (iBT) behandelt. Der primaere Endpunkt war die lokale Tumorkontrolle (LTC). Sekundaere Endpunkte umfassten die Zeit bis zum nicht mehr behandelbaren Progress (TTUP), die Zeit bis zum Progress (TTP), das Gesamtueberleben (OS) und die Sicherheit der Methode. In einer sekundaeren Analyse wurden Risikofaktoren

  20. Applicability and dosimetric impact of ultrasound-based preplanning in high-dose-rate brachytherapy of prostate cancer

    International Nuclear Information System (INIS)

    Aebersold, D.M.; Isaak, B.; Behrensmeier, F.; Kolotas, C.; Mini, R.; Greiner, R.H.; Thalmann, G.; Kranzbuehler, H.

    2004-01-01

    Background and purpose: analyses of permanent brachytherapy seed implants of the prostate have demonstrated that the use of a preplan may lead to a considerable decrease of dosimetric implant quality. The authors aimed to determine whether the same drawbacks of preplanning also apply to high-dose-rate (HDR) brachytherapy. Patients and methods: 15 patients who underwent two separate HDR brachytherapy implants in addition to external-beam radiation therapy for advanced prostate cancer were analyzed. A pretherapeutic transrectal ultrasound was performed in all patients to generate a preplan for the first brachytherapy implant. For the second brachytherapy, a subset of patients were treated by preplans based on the ultrasound from the first brachytherapy implant. Preplans were compared with the respective postplans assessing the following parameters: coverage index, minimum target dose, homogeneity index, and dose exposure of organs at risk. The prostate geometries (volume, width, height, length) were compared as well. Results: at the first brachytherapy, the matching between the preplan and actual implant geometry was sufficient in 47% of the patients, and the preplan could be applied. The dosimetric implant quality decreased considerably: the mean coverage differed by -0.11, the mean minimum target dose by -0.15, the mean homogeneity index by -0.09. The exposure of organs at risk was not substantially altered. At the second brachytherapy, all patients could be treated by the preplan; the differences between the implant quality parameters were less pronounced. The changes of prostate geometry between preplans and postplans were considerable, the differences in volume ranging from -8.0 to 13.8 cm 3 and in dimensions (width, height, length) from -1.1 to 1.0 cm. Conclusion: preplanning in HDR brachytherapy of the prostate is associated with a substantial decrease of dosimetric implant quality, when the preplan is based on a pretherapeutic ultrasound. The implant quality

  1. [Technique of intraoperative planning in prostatic brachytherapy with permanent implants of 125I or 103Pd].

    Science.gov (United States)

    Prada Gómez, Pedro José; Juan Rijo, Germán; Hevia Suarez, Miguel; Abascal García, José María; Abascal García, Ramón

    2002-12-01

    Prostatic brachytherapy with permanent 125I or 123Pd seeds implantation is a therapeutic option for organ-confined prostate cancer. We analyze the technique based on previous planning, our current intraoperative planning procedure and the reasons that moved us to introduce this change. Changes in prostate volume and spatial localization observed between previous planning and intraoperative images, and possible difficulties for seed implantation due to pubic arch interference are some of the reasons that induce us to change technique. Before the operation, we calculate the prostatic volume by transrectal ultrasound; with this information we determine the total implant activity following Wu's nomogram, and per-seed activity; therefore, it is an individual process for each patient. We perform a peripheral implant, placing 75-80% of the seeds within the peripheral prostatic zone, generally through 12-15 needles, the rest of the seeds are placed in the central prostatic zone using a maximum of 3-4 needles in high volume prostates. The day of intervention, after positioning and catheter insertion, volumetry is re-checked. Ultrasound images (from base to apex every 5 mm) are transferred to the planner were a suitable seed distribution is determined. Implantation is then performed placing all needles unloaded, and then intraoperative post-planning to allow us to check implant precision is performed after cistoscopically check that there is no urethral or bladder penetration by any needle. We finish with the insertion of seeds into the prostate. Total time for the procedure is around 90 minutes. Intraoperative planning is an additional step for the treatment of prostate cancer with permanent seeds brachytherapy, which avoids the disadvantages of previous planning and improves tumor inclusion in the ideal irradiation dose area, which will translate into better local disease control.

  2. An orthodontic device for retaining implanted radioactive sources during brachytherapy for cancer of the oral cavity

    International Nuclear Information System (INIS)

    Masuko, Noriko; Katsura, Kouji; Sugita, Tadashi; Sakai, Kunio; Sato, Katsurou; Kawana, Masahiro; Nonomura, Naobumi

    2000-01-01

    An orthodontic retainer was devised to keeping implanted radioactive sources in position and improve the quality of life during brachytherapy for cancer of the oral cavity. The retainer was used in 3 patients with oral cancer, one with cancer of the hard palate, one with cancer of the soft palate, and one with cancer of the floor of mouth, during brachytherapy using 198 Au grains and 137 Cs needles. These patients could speak freely. One with cancer of the hard palate could drink water and ingest semi-liquid food during treatment instead of nasal tube feeding. The plaster dental model obtained while making the retainer proved to be useful for training radiation oncologists. (author)

  3. Cosmetic results in early stage breast cancer patients with high-dose brachytherapy after conservative surgery

    International Nuclear Information System (INIS)

    Torres, Felipe; Pineda, Beatriz E

    2004-01-01

    Purpose: to reveal cosmetic results in patients at early stages of low risk breast cancer treated with partial accelerated radiotherapy using high dose rate brachytherapy. Methods and materials: from March 2001 to July 2003,14 stages l and ll breast cancer patients were treated at the Colombian national cancer institute in Bogota with conservative surgery and radiotherapy upon the tumor bed (partial accelerated radiotherapy), using interstitial implants with iridium 192 (high dose rate brachytherapy) with a dose of 32 Gys, over 4 days, at 8 fractions twice a day. Results: with an average follow up of 17.7 months, good cosmetic results were found among 71.4 % of patients and excellent results among 14.3% of patients, furthermore none of the patients neither local nor regional or distant relapses. Conclusion: among patients who suffer from breast cancer at early stages, it showed is possible to apply partial accelerated radiotherapy upon the tumor bed with high doses over 4 days with good to excellent cosmetic results

  4. Simultaneous treatment of tongue cancer with interstitial brachytherapy and bleomycin

    International Nuclear Information System (INIS)

    Watarai, Jiro; Itagaki, Takatomo; Yamaguchi, Kouichi

    1983-01-01

    During a period of 5 years, from 1977 to 1982, twenty five patients with tongue cancer were treated by radium needle implantation and bleomycin at Yamagata University Hospital. In this paper, authors analysed seventeen patients followed over two years. All had biopsy proven squamous cell carcinoma. According to the TNM system (UICC, 1978), primary tumor was classified into 4 cases of T1, 8 cases of T2 and 5 cases of T3. The main purpose of this study was to obtain a high local control rate and reduce subsequent regional lymphnode metastasis. Our curative treatment method was simultaneous combination of 70 Gy of brachy-therapy and 40 mg of bleomycin. The results of this study were as follows: 1. A control rate in the primary lesion was 91% (10/11) in survivors having survived more than 2 years. 2. Radioosteonecrosis of mandible was found in 6% (1/17) and transient ulcer formation in the primary site was observed in 35% (6/17) of patients treated. However, all patients were cured by conservative treatment. 3. This treatment method did not reduce subsequent lymph node metastasis. (author)

  5. MO-B-BRC-01: Introduction [Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Prisciandaro, J. [University of Michigan (United States)

    2016-06-15

    Brachytherapy has proven to be an effective treatment option for prostate cancer. Initially, prostate brachytherapy was delivered through permanently implanted low dose rate (LDR) radioactive sources; however, high dose rate (HDR) temporary brachytherapy for prostate cancer is gaining popularity. Needle insertion during prostate brachytherapy is most commonly performed under ultrasound (U/S) guidance; however, treatment planning may be performed utilizing several imaging modalities either in an intra- or post-operative setting. During intra-operative prostate HDR, the needles are imaged during implantation, and planning may be performed in real time. At present, the most common imaging modality utilized for intra-operative prostate HDR is U/S. Alternatively, in the post-operative setting, following needle implantation, patients may be simulated with computed tomography (CT) or magnetic resonance imaging (MRI). Each imaging modality and workflow provides its share of benefits and limitations. Prostate HDR has been adopted in a number of cancer centers across the nation. In this educational session, we will explore the role of U/S, CT, and MRI in HDR prostate brachytherapy. Example workflows and operational details will be shared, and we will discuss how to establish a prostate HDR program in a clinical setting. Learning Objectives: Review prostate HDR techniques based on the imaging modality Discuss the challenges and pitfalls introduced by the three imagebased options for prostate HDR brachytherapy Review the QA process and learn about the development of clinical workflows for these imaging options at different institutions.

  6. CT-image-based conformal brachytherapy of breast cancer. The significance of semi-3-D and 3-D treatment planning.

    Science.gov (United States)

    Polgár, C; Major, T; Somogyi, A; Takácsi-Nagy, Z; Mangel, L C; Forrai, G; Sulyok, Z; Fodor, J; Németh, G

    2000-03-01

    To compare the conventional 2-D, the simulator-guided semi-3-D and the recently developed CT-guided 3-D brachytherapy treatment planning in the interstitial radiotherapy of breast cancer. In 103 patients with T1-2, N0-1 breast cancer the tumor bed was clipped during breast conserving surgery. Fifty-two of them received boost brachytherapy after 46 to 50 Gy teletherapy and 51 patients were treated with brachytherapy alone via flexible implant tubes. Single, double and triple plane implant was used in 6, 89 and 8 cases, respectively. The dose of boost brachytherapy and sole brachytherapy prescribed to dose reference points was 3 times 4.75 Gy and 7 times 5.2 Gy, respectively. The positions of dose reference points varied according to the level (2-D, semi-3-D and 3-D) of treatment planning performed. The treatment planning was based on the 3-D reconstruction of the surgical clips, implant tubes and skin points. In all cases the implantations were planned with a semi-3-D technique aided by simulator. In 10 cases a recently developed CT-guided 3-D planning system was used. The semi-3-D and 3-D treatment plans were compared to hypothetical 2-D plans using dose-volume histograms and dose non-uniformity ratios. The values of mean central dose, mean skin dose, minimal clip dose, proportion of underdosaged clips and mean target surface dose were evaluated. The accuracy of tumor bed localization and the conformity of planning target volume and treated volume were also analyzed in each technique. With the help of conformal semi-3-D and 3-D brachytherapy planning we could define reference dose points, active source positions and dwell times individually. This technique decreased the mean skin dose with 22.2% and reduced the possibility of geographical miss. We could achieve the best conformity between the planning target volume and the treated volume with the CT-image based 3-D treatment planning, at the cost of worse dose homogeneity. The mean treated volume was reduced by 25

  7. Stereotactic iodine-125 brachytherapy for brain tumors: temporary versus permanent implantation

    Directory of Open Access Journals (Sweden)

    Ruge Maximilian I

    2012-06-01

    Full Text Available Abstract Stereotactic brachytherapy (SBT has been described in several publications as an effective, minimal invasive and safe highly focal treatment option in selected patients with well circumscribed brain tumors 40 cGy/h in combination with adjuvant external beam radiation and/or chemotherapy for the treatment of malignant gliomas and metastases resulted in increased rates of radiation induced adverse tissue changes requiring surgical intervention. Vice versa, such effects have been only minimally observed in numerous studies applying low dose rate (LDR regiments (3–8 cGy/h for low grade gliomas, metastases and other rare indications. Besides these observations, there are, however, no data available directly comparing the long term incidences of tissue changes after HDR and LDR and there is, furthermore, no evidence regarding a difference between temporary or permanent LDR implantation schemes. Thus, recommendations for effective and safe implantation schemes have to be investigated and compared in future studies.

  8. Results of the intestitial brachytherapy and of the combination external radiation-brachytherapy in 150 patients with carcinoma of the oral tongue and floor of the mouth

    Energy Technology Data Exchange (ETDEWEB)

    Sannazzari, G L; Negri, G L; Ozzello, F

    1986-01-01

    The authors report their experience on the treatment of carcinoma of the oral tongue and floor of the mouth with interstitial brachytherapy, alone or in conbination with external irradiation. One hundred and fifty patients were treated; among these, 116 with brachytherapy alone, 34 with combined treatment. The five years local control in those patients treated with brachytherapy alone was 72.5% in T1, 61.2% in T2 and 35% in T3; in those patients treated with external irradiation and brachytherapy the global five years control was 42.5%. The global five years survival was 64% in the patients treated with brachytherapy alone 48% in the patients treated with combined therapy. 42 refs.

  9. Results of the intestitial brachytherapy and of the combination external radiation-brachytherapy in 150 patients with carcinoma of the oral tongue and floor of the mouth

    International Nuclear Information System (INIS)

    Sannazzari, G.L.; Negri, G.L.; Ozzello, F.

    1986-01-01

    The authors report their experience on the treatment of carcinoma of the oral tongue and floor of the mouth with interstitial brachytherapy, alone or in conbination with external irradiation. One hundred and fifty patients were treated; among these, 116 with brachytherapy alone, 34 with combined treatment. The five years local control in those patients treated with brachytherapy alone was 72.5% in T1, 61.2% in T2 and 35% in T3; in those patients treated with external irradiation and brachytherapy the global five years control was 42.5%. The global five years survival was 64% in the patients treated with brachytherapy alone 48% in the patients treated with combined therapy

  10. SU-F-J-157: Effect of Contouring Uncertainty in Post Implant Dosimetry of Low-Dose-Rate Prostate Permanent Seed Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Mashouf, S; Merino, T; Ravi, A; Morton, G; Song, W [Sunnybrook Health Sciences Center, Odette Cancer Centre, Toronto, ON (Canada); University of Toronto, Dept. of Radiation Oncology, Toronto, ON (Canada); Safigholi, H; Soliman, A [Sunnybrook Research Institute, Toronto, ON (Canada)

    2016-06-15

    Purpose: There is strong evidence relating post-implant dosimetry for low-dose-rate (LDR) prostate seed brachytherapy to local control rates. The delineation of the prostate on CT images, however, represents a challenge due to the lack of soft tissue contrast in order to identify the prostate borders. This study aims at quantifying the sensitivity of clinically relevant dosimetric parameters to uncertainty in the contouring of prostate. Methods: CT images, post-op plans and contours of a cohort of patients (n=43) (low risk=55.8%, intermediate risk=39.5%, high risk=4.7%), who had received prostate seed brachytherapy, were imported into MIM Symphony treatment planning system. The prostate contours in post-implant CT images were expanded/contracted uniformly for margins of ±1.00 mm, ±2.00 mm, ±3.00 mm, ±4.00 mm and ±5.00 mm. The values for V100 and D90 were extracted from Dose Volume Histograms for each contour and compared. Results: Significant changes were observed in the values of D90 and V100 as well as the number of suboptimal plans for expansion or contraction margins of only few millimeters. Evaluation of coverage based on D90 was found to be less sensitive to expansion errors compared to V100. D90 led to a lower number of implants incorrectly identified with insufficient coverage for expanded contours which increases the accuracy of post-implant QA using CT images compared to V100. Conclusion: In order to establish a successful post implant QA for LDR prostate seed brachytherapy, it is necessary to identify the low and high thresholds of important dose metrics of the target volume such as D90 and V100. Since these parameters are sensitive to target volume definition, accurate identification of prostate borders would help to improve accuracy and predictive value of the post-implant QA process. In this respect, use of imaging modalities such as MRI where prostate is well delineated should prove useful.

  11. SU-F-J-157: Effect of Contouring Uncertainty in Post Implant Dosimetry of Low-Dose-Rate Prostate Permanent Seed Brachytherapy

    International Nuclear Information System (INIS)

    Mashouf, S; Merino, T; Ravi, A; Morton, G; Song, W; Safigholi, H; Soliman, A

    2016-01-01

    Purpose: There is strong evidence relating post-implant dosimetry for low-dose-rate (LDR) prostate seed brachytherapy to local control rates. The delineation of the prostate on CT images, however, represents a challenge due to the lack of soft tissue contrast in order to identify the prostate borders. This study aims at quantifying the sensitivity of clinically relevant dosimetric parameters to uncertainty in the contouring of prostate. Methods: CT images, post-op plans and contours of a cohort of patients (n=43) (low risk=55.8%, intermediate risk=39.5%, high risk=4.7%), who had received prostate seed brachytherapy, were imported into MIM Symphony treatment planning system. The prostate contours in post-implant CT images were expanded/contracted uniformly for margins of ±1.00 mm, ±2.00 mm, ±3.00 mm, ±4.00 mm and ±5.00 mm. The values for V100 and D90 were extracted from Dose Volume Histograms for each contour and compared. Results: Significant changes were observed in the values of D90 and V100 as well as the number of suboptimal plans for expansion or contraction margins of only few millimeters. Evaluation of coverage based on D90 was found to be less sensitive to expansion errors compared to V100. D90 led to a lower number of implants incorrectly identified with insufficient coverage for expanded contours which increases the accuracy of post-implant QA using CT images compared to V100. Conclusion: In order to establish a successful post implant QA for LDR prostate seed brachytherapy, it is necessary to identify the low and high thresholds of important dose metrics of the target volume such as D90 and V100. Since these parameters are sensitive to target volume definition, accurate identification of prostate borders would help to improve accuracy and predictive value of the post-implant QA process. In this respect, use of imaging modalities such as MRI where prostate is well delineated should prove useful.

  12. Accelerated partial-breast irradiation with interstitial implants. The clinical relevance of the calculation of skin doses

    International Nuclear Information System (INIS)

    Ott, O.J.; Lotter, M.; Sauer, R.; Strnad, V.

    2007-01-01

    Purpose: To describe relative skin dose estimations and their impact on cosmetic outcome in interstitial multicatheter accelerated partial-breast irradiation (APBI). Patients and Methods: Between April 2001 and January 2005, 105 consecutive patients with early breast cancer were recruited in Erlangen, Germany, for this substudy of the German-Austrian APBI phase II trial. 51% (54/105) received pulsed-dose-rate (PDR), and 49% (51/105) high-dose-rate (HDR) brachytherapy. Prescribed reference dose for HDR brachytherapy was 32 Gy in eight fractions of 4 Gy, twice daily. Prescribed reference dose in PDR brachytherapy was 49.8 Gy in 83 consecutive fractions of 0.6 Gy every hour. Total treatment time was 3-4 days. With a wire cross on the skin surface during the brachytherapy-planning procedure the minimal, mean and maximal relative skin doses (SD min% , SD max% , SD mean% ) were recorded. Endpoint of this evaluation was the cosmetic outcome in relation to the relative skin doses. Results: Median follow-up time was 38 months (range, 19-65 months). Cosmetic results for all patients were excellent in 57% (60/105), good in 36% (38/105), and fair in 7% (7/105). The SD min% (27.0% vs. 31.7%; p = 0.032), SD mean% (34.2% vs. 38.1%; p 0.008), and SD max% (38.2% vs. 46.4%; p 0.003) were significantly lower for patients with excellent cosmetic outcome compared to patients with a suboptimal outcome. SD mean% (37.6% vs. 34.2%; p = 0.026) and SD max% (45.4% vs. 38.2%; p = 0.008) were significantly higher for patients with good cosmetic outcome compared with the patients with excellent results. Conclusion: The appraisal of skin doses has been shown to be relevant to the achievement of excellent cosmetic outcome. Further investigations are necessary, especially on the basis of CT-based brachytherapy planning, to further improve the treatment results of multicatheter APBI. (orig.)

  13. Severe rectal complications after prostate brachytherapy

    International Nuclear Information System (INIS)

    Wallner, Kent; Sutlief, Stephen; Bergsagel, Carl; Merrick, Gregory S.

    2015-01-01

    Purpose: Some investigators have reported severe rectal complications after brachytherapy. Due to the low number of such events, their relationship to dosimetric parameters has not been well characterized. Methods and materials: A total of 3126 patients were treated with low dose rate brachytherapy from 1998 through 2010. 2464 had implant alone, and 313 had implant preceded by 44–46 Gy supplemental external beam radiation (EBRT). Post-implant dosimetry was based on a CT scan obtained on the day of implant, generally within 30 min of the procedure. Every patient’s record was reviewed for occurrence of rectal complications. Results: Eight of 2464 patients (0.32%) treated with brachytherapy alone developed a radiation-related rectal fistula. Average prostatic and rectal dose parameters were moderately higher for fistula patients than for patients without a severe rectal complication. For instance, the average R100 was 1.2 ± 0.75 cc for fistula patients, versus 0.37 ± 0.88 cc for non-fistula patients. However, the fistula patients’ values were well within the range of values for patients without a rectal complication. Four patients had some attempt at repair or reconstruction, but long-term functional outcomes were not favorable. Conclusions: Rectal fistulas are a very uncommon potential complication of prostate brachytherapy, which can occur even in the setting of acceptable day 0 rectal doses. Their occurrence is not easily explained by standard dosimetric or clinical factors

  14. Evaluation of radiation dose on people adjacent to implant patients during brachytherapy for prostate cancer using {sup 192}Ir

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Jung Hoon; Ko, Seong Jin; Kang, Se Sik; Kim, Chang Soo [Catholic University, Busan (Korea, Republic of)

    2009-10-15

    The incidence of prostate cancer is rapidly increasing due to aging of the population and westernization of dietary habits, etc. As a result, the frequency of prostate cancer has become the fifth highest among all male cancers and the first among urological cancers. Brachytherapy is commonly used for locally progressing prostate cancers. Since the mid 1980s, therapies using radio-isotopes, such as low-invasive {sup 125}I, {sup 103}Pd and {sup 192}Ir, have been widely performed in the U.S. and Europe. However, brachytherapy involves implanting radio-isotopes into the human body which is of concern because it may expose the health care professionals administering the therapy to unnecessary radiation. Accordingly, this study intends to predict the radiation dose that people adjacent to patients implanted with a radio-isotope are exposed to during prostate cancer radiation therapy by using a mathematical anthropomorphic phantom and {sup 192}Ir.

  15. 10 CFR 35.2406 - Records of brachytherapy source accountability.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 1 2010-01-01 2010-01-01 false Records of brachytherapy source accountability. 35.2406... Records of brachytherapy source accountability. (a) A licensee shall maintain a record of brachytherapy source accountability required by § 35.406 for 3 years. (b) For temporary implants, the record must...

  16. Preliminary results of a phase I/II study of HDR brachytherapy alone for T1/T2 breast cancer

    International Nuclear Information System (INIS)

    Wazer, David E.; Berle, Lisa; Graham, Roger; Chung, Maureen; Rothschild, Janice; Graves, Theresa; Cady, Blake; Ulin, Kenneth; Ruthazer, Robin; DiPetrillo, Thomas A.

    2002-01-01

    Purpose: To investigate the feasibility, toxicity, cosmetic outcome, and local control of high-dose-rate (HDR) brachytherapy alone without whole breast external beam irradiation for early-stage breast carcinoma. Methods and Materials: Between June 1997 and August 1999, 32 women diagnosed with a total of 33 AJCC Stage I/II breast carcinomas underwent surgical breast excision and postoperative irradiation using HDR brachytherapy interstitial implantation as part of a multi-institutional clinical Phase I/II protocol. Eligible patients included those with T1, T2, N0, N1 (≤3 nodes positive), and M0 tumors of nonlobular histologic features with negative surgical margins, no extracapsular lymph node extension, and a negative postexcision mammogram. Brachytherapy catheters were placed at the initial excision, reexcision, or either sentinel or full-axillary sampling. Direct visualization, surgical clips, and ultrasound and/or CT scan assisted in the delineation of the target volume, defined as the excision cavity plus a 2-cm margin. High-activity 192 Ir (3-10 Ci) was used to deliver 340 cGy/fraction, 2 fractions/d, for 5 consecutive days, to a total dose of 34 Gy to the target volume. Source position and dwell times were calculated using standard volume optimization techniques. Results: The median follow-up of all patients was 33 months, and the mean patient age was 63 years. The mean tumor size was 1.3 cm, and 55% had an extensive intraductal component. Three patients had positive axillary nodes. Two patients experienced moderate perioperative pain that required narcotic analgesics. No peri- or postoperative infections occurred. No wound healing problems and no significant skin reactions related to the implant developed. The Radiation Therapy Oncology Group late radiation morbidity scoring scheme was applied to the entire 33-case cohort. In the assessment of the skin, 30 cases were Grade 0-1 and 3 cases were Grade 2. Subcutaneous toxicity was scored as 11 patients with

  17. Perioperative high dose rate (HDR brachytherapy in unresectable locally advanced pancreatic tumors

    Directory of Open Access Journals (Sweden)

    Brygida Białas

    2011-07-01

    Full Text Available Purpose: The aim of the study was to present an original technique of catheter implantation for perioperative HDR-Ir192 brachytherapy in patients after palliative operations of unresectable locally advanced pancreatic tumors and to estimate the influence of perioperative HDR-Ir192 brachytherapy on pain relief in terminal pancreatic cancer patients. Material and methods: Eight patients with pancreatic tumors located in the head of pancreas underwent palliative operations with the use of HDR-Ir192 brachytherapy. All patients qualified for surgery reported pain of high intensity and had received narcotic painkillers prior to operation. During the last phase of the surgery, the Nucletron® catheters were implanted in patients to prepare them for later perioperative brachytherapy. Since the 6th day after surgery HDR brachytherapy was performed. Before each brachytherapy fraction the location of implants were checked using fluoroscopy. A fractional dose was 5 Gy and a total dose was 20 Gy in the area of radiation. A comparative study of two groups of patients (with and without brachytherapy with stage III pancreatic cancer according to the TNM scale was taken in consideration. Results and Conclusions: The authors claim that the modification of catheter implantation using specially designed cannula, facilitates the process of inserting the catheter into the tumor, shortens the time needed for the procedure, and reduces the risk of complications. Mean survival time was 5.7 months. In the group of performed brachytherapy, the mean survival time was 6.7 months, while in the group of no brachytherapy performed – 4.4 months. In the group of brachytherapy, only one patient increased the dose of painkillers in the last month of his life. Remaining patients took constant doses of medicines. Perioperative HDR-Ir192 brachytherapy could be considered as a practical application of adjuvant therapy for pain relief in patients with an advanced pancreatic cancer.

  18. Urethral toxicity after LDR brachytherapy: experience in Japan.

    Science.gov (United States)

    Tanaka, Nobumichi; Asakawa, Isao; Hasegawa, Masatoshi; Fujimoto, Kiyohide

    2015-01-01

    Urinary toxicity is common after low-dose-rate (LDR) brachytherapy, and the resolution of urinary toxicity is a concern. In particular, urinary frequency is the most common adverse event among the urinary toxicities. We have previously reported that approximately 70% of patients experience urinary frequency during the first 6 months after seed implantation. Most urinary adverse events were classified as Grade 1, and Grade 2 or higher adverse events were rare. The incidence of urinary retention was approximately 2-4%. A high International Prostate Symptom Score before seed implantation was an independent predictor of acute urinary toxicity of Grade 2 or higher. Several previous reports from the United States also supported this trend. In Japan, LDR brachytherapy was legally approved in 2003. A nationwide prospective cohort study entitled Japanese Prostate Cancer Outcome Study of Permanent Iodine-125 Seed Implantation was initiated in July 2005. It is an important issue to limit urinary toxicities in patients who undergo LDR brachytherapy. Copyright © 2015 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  19. Accelerated partial breast irradiation: An analysis of variables associated with late toxicity and long-term cosmetic outcome after high-dose-rate interstitial brachytherapy

    International Nuclear Information System (INIS)

    Wazer, David E.; Kaufman, Seth; Cuttino, Laurie; Di Petrillo, Thomas; Arthur, Douglas W.

    2006-01-01

    Purpose: To perform a detailed analysis of variables associated with late tissue effects of high-dose-rate (HDR) interstitial brachytherapy accelerated partial breast irradiation (APBI) in a large cohort of patients with prolonged follow-up. Methods and Materials: Beginning in 1995, 75 women with Stage I/II breast cancer were enrolled in identical institutional trials evaluating APBI as monotherapy after lumpectomy. Patients eligible included those with T1-2, N0-1 (≤3 nodes positive), M0 tumors of nonlobular histology with negative surgical margins, no extracapsular nodal extension, and negative results on postexcision mammogram. All patients underwent surgical excision and postoperative irradiation with HDR interstitial brachytherapy. The planning target volume was defined as the excision cavity plus a 2-cm margin. Treatment was delivered with a high-activity Ir-192 source at 3.4 Gy per fraction twice daily for 5 days to a total dose of 34 Gy. Dosimetric analyses were performed with three-dimensional postimplant dose and volume reconstructions. All patients were evaluated at 3-6-month intervals and assessed with a standardized cosmetic rating scale and according to Radiation Therapy Oncology Group late normal tissue toxicity scoring criteria. Clinical and therapy-related features were analyzed for their relationship to cosmetic outcome and toxicity rating. Clinical features analyzed included age, volume of resection, history of diabetes or hypertension, extent of axillary surgery, and systemic therapies. Therapy-related features analyzed included volume of tissue encompassed by the 100%, 150%, and 200% isodose lines (V100, V150, and V200, respectively), the dose homogeneity index (DHI), number of source dwell positions, and planar separation. Results: The median follow-up of all patients was 73 months (range, 43-118 months). The cosmetic outcome at last follow-up was rated as excellent, good, and fair/poor in 67%, 24%, and 9% of patients, respectively

  20. Accelerated partial breast irradiation utilizing balloon brachytherapy techniques

    International Nuclear Information System (INIS)

    Strauss, Jonathan B.; Dickler, Adam

    2009-01-01

    To overcome the barriers to BCT, methods of PBI in the setting of breast conservation have been explored. The method of PBI with the longest published follow-up is multi-catheter interstitial brachytherapy. Balloon-based brachytherapy with the MammoSite brachytherapy applicator was designed to simplify the brachytherapy procedure for PBI, enhance the reproducibility of the dosimetry, and improve patient comfort. The rates of local recurrence following PBI with the MammoSite applicator have been low, but there are few published reports and follow-up has been relatively short. The cosmetic outcomes and toxicity of MammoSite PBI are comparable to those seen after multicatheter-based PBI. Additional methods of balloon brachytherapy, including Xoft and SenoRx Contura have been developed. Finally, long-term follow-up after PBI is important for the welfare of individual patients and in order to establish the efficacy, late toxicity and cosmetic outcomes of this technique.

  1. Radiobiology for eye plaque brachytherapy and evaluation of implant duration and radionuclide choice using an objective function

    International Nuclear Information System (INIS)

    Gagne, Nolan L.; Leonard, Kara L.; Rivard, Mark J.

    2012-01-01

    Purpose: Clinical optimization of Collaborative Ocular Melanoma Study (COMS) eye plaque brachytherapy is currently limited to tumor coverage, consensus prescription dosage, and dose calculations to ocular structures. The biologically effective dose (BED) of temporary brachytherapy treatments is a function of both chosen radionuclide R and implant duration T. This study endeavored to evaluate BED delivered to the tumor volume and surrounding ocular structures as a function of plaque position P, prescription dose, R, and T. Methods: Plaque-heterogeneity-corrected dose distributions were generated with MCNP5 for the range of currently available COMS plaques loaded with sources using three available low-energy radionuclides. These physical dose distributions were imported into the PINNACLE 3 treatment planning system using the TG-43 hybrid technique and used to generate dose volume histograms for a T = 7 day implant within a reference eye geometry including the ciliary body, cornea, eyelid, foveola, lacrimal gland, lens, optic disc, optic nerve, retina, and tumor at eight standard treatment positions. The equation of Dale and Jones was employed to create biologically effective dose volume histograms (BEDVHs), allowing for BED volumetric analysis of all ROIs. Isobiologically effective prescription doses were calculated for T = 5 days down to 0.01 days, with BEDVHs subsequently generated for all ROIs using correspondingly reduced prescription doses. Objective functions were created to evaluate the BEDVHs as a function of R and T. These objective functions are mathematically accessible and sufficiently general to be applied to temporary or permanent brachytherapy implants for a variety of disease sites. Results: Reducing T from 7 to 0.01 days for a 10 mm plaque produced an average BED benefit of 26%, 20%, and 17% for 103 Pd, 125 I, and 131 Cs, respectively, for all P; 16 and 22 mm plaque results were more position-dependent. 103 Pd produced a 16%–35% BED benefit over

  2. Use of brachytherapy with permanent implants of iodine-125 in localized prostate cancer

    International Nuclear Information System (INIS)

    Bladou, F.; Serment, G.; Salem, N.; Simonian, M.; Rosello, R.; Ternier, F.

    2002-01-01

    Approximately 15,000 cases of early stage prostate cancer T1 and T2 are diagnosed every year in France by testing for PSA and performing prostatic biopsies. The treatment of these localized forms is based in most cases on radical prostatectomy or nn external beam radiotherapy. Although the ontological results obtained by these two therapeutic methods are satisfactory and equivalent in the long term, the side effects can be important. For a number of years, trans-perineal brachytherapy using permanent implants of iodine -125 or palladium-103 has proved itself as an alternative therapy with equivalent medium to long-term results. The low urinary, digestive and sexual side effects of prostate brachytherapy are important reasons for the enthusiasm among patients and the medical community for this therapy and the growing number of applications and centres which practice it. In September 1998 we started the prostate brachytherapy programmes- in Marseilles with close collaboration between the department of urology of the Hopital Salvator, and the departments of radiotherapy, medical imaging and medical physics of the Institut Paoli-Calmettes. To date, around 250 patients with localized adenocarcinoma of the prostate have benefited from this alternative therapy in our centre. Preliminary results, with a 3 year-follow-up, are comparable to results published in the literature by pioneer teams. (authors)

  3. Brachytherapy of endometrial cancers

    International Nuclear Information System (INIS)

    Peiffert, D.; Hoffstetter, S.; Charra-Brunaud, C.

    2003-01-01

    Endometrial adenocarcinomas rank third as tumoral sites en France. The tumors are confined to the uterus in 80% of the cases. Brachytherapy has a large place in the therapeutic strategy. The gold standard treatment remains extra-fascial hysterectomy with bilateral annexiectomy and bilateral internal iliac lymph node dissection. However, after surgery alone, the rate of locoregional relapses reaches 4-20%, which is reduced to 0-5% after postoperative brachytherapy of the vaginal cuff. This postoperative brachytherapy is delivered as outpatients treatment, by 3 or 4 fractions, at high dose rate. The utero-vaginal preoperative brachytherapy remains well adapted to the tumors which involve the uterine cervix. Patients presenting a localized tumor but not operable for general reasons (< 10%) can be treated with success by exclusive irradiation, which associates a pelvic irradiation followed by an utero-vaginal brachytherapy. A high local control of about 80-90% is obtained, a little lower than surgery, with a higher risk of late complications. Last but not least, local relapses in the vaginal cuff, or in the perimeatic area, can be treated by interstitial salvage brachytherapy, associated if possible with external beam irradiation. The local control is reached in half of the patients, but metastatic dissemination is frequent. We conclude that brachytherapy has a major role in the treatment of endometrial adenocarcinomas, in combination with surgery, or with external beam irradiation for not operable patients or in case of local relapses. It should use new technologies now available including computerized after-loaders and 3D dose calculation. (authors)

  4. Evaluation of time, attendance of medical staff, and resources during interstitial brachytherapy for prostate cancer. DEGRO-QUIRO trial

    International Nuclear Information System (INIS)

    Tselis, N.; Zamboglou, N.; Maurer, U.; Popp, W.; Sack, H.

    2014-01-01

    The German Society of Radiation Oncology initiated a multicenter trial to evaluate core processes and subprocesses of radiotherapy by prospective evaluation of all important procedures in the most frequent malignancies treated by radiation therapy. The aim of this analysis was to assess the required resources for interstitial high-dose-rate (HDR) and low-dose-rate (LDR) prostate brachytherapy (BRT) based on actual time measurements regarding allocation of personnel and room occupation needed for specific procedures. Two radiotherapy centers (community hospital of Offenbach am Main and community hospital of Eschweiler) participated in this prospective study. Working time of the different occupational groups and room occupancies for the workflow of prostate BRT were recorded and methodically assessed during a 3-month period. For HDR and LDR BRT, a total of 560 and 92 measurements, respectively, were documented. The time needed for treatment preplanning was median 24 min for HDR (n=112 measurements) and 6 min for LDR BRT (n=21). Catheter implantation with intraoperative HDR real-time planning (n=112), postimplantation HDR treatment planning (n=112), and remotely controlled HDR afterloading irradiation (n=112) required median 25, 39, and 50 min, respectively. For LDR real-time planning (n=39) and LDR treatment postplanning (n=32), the assessed median duration was 91 and 11 min, respectively. Room occupancy and overall mean medical staff times were 194 and 910 min respectively, for HDR, and 113 and 371 min, respectively, for LDR BRT. In this prospective analysis, the resource requirements for the application of HDR and LDR BRT of prostate cancer were assessed methodically and are presented for first time. (orig.)

  5. Evaluation of time, attendance of medical staff, and resources during interstitial brachytherapy for prostate cancer. DEGRO-QUIRO trial

    Energy Technology Data Exchange (ETDEWEB)

    Tselis, N.; Zamboglou, N. [Sana Klinikum Offenbach, Department of Radiation Oncology, Offenbach am Main (Germany); Maurer, U. [St.-Antonius-Hospital, Strahlentherapie, Eschweiler (Germany); Popp, W. [Prime Networks AG, Basel (Switzerland); Sack, H. [University of Essen, Department of Radiation Oncology, Essen (Germany)

    2014-04-15

    The German Society of Radiation Oncology initiated a multicenter trial to evaluate core processes and subprocesses of radiotherapy by prospective evaluation of all important procedures in the most frequent malignancies treated by radiation therapy. The aim of this analysis was to assess the required resources for interstitial high-dose-rate (HDR) and low-dose-rate (LDR) prostate brachytherapy (BRT) based on actual time measurements regarding allocation of personnel and room occupation needed for specific procedures. Two radiotherapy centers (community hospital of Offenbach am Main and community hospital of Eschweiler) participated in this prospective study. Working time of the different occupational groups and room occupancies for the workflow of prostate BRT were recorded and methodically assessed during a 3-month period. For HDR and LDR BRT, a total of 560 and 92 measurements, respectively, were documented. The time needed for treatment preplanning was median 24 min for HDR (n=112 measurements) and 6 min for LDR BRT (n=21). Catheter implantation with intraoperative HDR real-time planning (n=112), postimplantation HDR treatment planning (n=112), and remotely controlled HDR afterloading irradiation (n=112) required median 25, 39, and 50 min, respectively. For LDR real-time planning (n=39) and LDR treatment postplanning (n=32), the assessed median duration was 91 and 11 min, respectively. Room occupancy and overall mean medical staff times were 194 and 910 min respectively, for HDR, and 113 and 371 min, respectively, for LDR BRT. In this prospective analysis, the resource requirements for the application of HDR and LDR BRT of prostate cancer were assessed methodically and are presented for first time. (orig.)

  6. The theoretical basis and clinical methodology for stereotactic interstitial brain tumor irradiation using iododeoxyuridine as a radiation sensitizer and samarium-145 as a brachytherapy source

    International Nuclear Information System (INIS)

    Goodman, J.H.; Gahbauer, R.A.; Kanellitsas, C.; Clendenon, N.R.; Laster, B.H.; Fairchild, R.G.

    1989-01-01

    High grade astrocytomas have proven resistant to all conventional therapy. A technique to produce radiation enhancement during interstitial brain tumor irradiation by using a radiation sensitizer (IdUrd) and by stimulation of Auger electron cascades through absorption of low energy photons in iodine (Photon activation) is described. Clinical studies using IdUrd, 192 Ir as a brachytherapy source, and external radiation have produced promising results. Substituting samarium-145 for 192 Ir in this protocol is expected to produce enhanced results. 15 refs

  7. The Curie–Da Vinci Connection: 5-Years' Experience With Laparoscopic (Robot-Assisted) Implantation for High-Dose-Rate Brachytherapy of Solitary T2 Bladder Tumors

    Energy Technology Data Exchange (ETDEWEB)

    Steen-Banasik, Elzbieta M. van der, E-mail: E.vanderSteen-Banasik@radiotherapiegroep.nl [Radiotherapiegroep, Arnhem (Netherlands); Smits, Geert A.H.J. [Department of Urology, Rijnstate Hospital, Arnhem (Netherlands); Oosterveld, Bernard J.; Janssen, Theo; Visser, Andries G. [Radiotherapiegroep, Arnhem (Netherlands)

    2016-08-01

    Purpose: To report experience and early results of laparoscopic implantation for interstitial brachytherapy (BT) of solitary bladder tumors and the feasibility of a high-dose-rate (HDR) schedule. Methods and Materials: From December 2009 to April 2015, 57 patients with a T2 solitary bladder tumor were treated in Arnhem with transurethral bladder resection followed by external beam irradiation, applied to the bladder and regional iliac lymph nodes, 40 Gy in 20 fractions, 5 fractions per week, and within 1 week interstitial HDR BT, in selected cases combined with partial cystectomy and lymph node dissection. The BT catheters were placed via a transabdominal approach with robotic assistance from a Da Vinci robot after a successful initial experience with a nonrobotic laparoscopic approach. The fraction schedule for HDR was 10 fractions of 2.5 Gy, 3 fractions per day. This was calculated to be equivalent to a reference low-dose-rate schedule of 30 Gy in 60 hours. Data for oncologic outcomes and toxicity (Common Toxicity Criteria version 4) were prospectively collected. Results: These modifications resulted in an average postoperative hospitalization of 6 days, minimal blood loss, and no wound healing problems. Two patients had severe acute toxicity: 1 pulmonary embolism grade 4 and 1 cardiac death. Late toxicity was mild (n=2 urogenital grade 3 toxicity). The median follow-up was 2 years. Using cumulative incidence competing risk analysis, the 2-year overall, disease-free, and disease-specific survival and local control rates were 59%, 71%, 87%, and 82%, respectively. Conclusions: The benefits of minimally invasive surgery for implantation of BT catheters and the feasibility of HDR BT in bladder cancer are documented. The patient outcome and adverse events are comparable to the best results published for a bladder-sparing approach.

  8. Dosimetric results in implant and post-implant and low rate in brachytherapy prostate cancer with loose seeds and attached; Resultados dosimetricos en el implante y post-impante en braquiterapia de baja tasa en cancer de prostata con semillas sueltas y unidas

    Energy Technology Data Exchange (ETDEWEB)

    Juan-Senabre, X. J.; Albert Antequera, M.; Lopez-Tarjuelo, J.; Santos Serra, A.; Perez-Mestre, M.; Sanchez Iglesias, A. L.; Conde Moreno, A. J.; Gonzalez Vidal, V.; Beltran Persiva, J.; Muelas Soria, R.; Ferrer Albiach, C.

    2015-07-01

    The objective is determine differences dosimetry statistics on the dosimetry of the implant and post-implant in brachytherapy of low rate with implants permanent in prostate using seed of 125-I loose and attached Both in lives and in the post-prostatic plans dosimetric coverage is good and restrictions in urethra and rectum for both groups of patients are met. Not migrating with joined is evident, as well as better dosimetric homogeneity. (Author)

  9. Perioperative interstitial brachytherapy for recurrent keloid scars; La curietherapie interstitielle perioperatoire dans le traitement des cicatrices cheloides recidivantes

    Energy Technology Data Exchange (ETDEWEB)

    Rio, E; Bardet, E; Peuvrel, P [CRLCC Nantes-Atlantique, Service de radiotherapie, 44 - Nantes-St-Herblain (France); Martinet, L; Perrot, P [CHU Hotel-Dieu, Service de chirurgie plastique, 44 - Nantes (France); Baraer, F; Loirat, Y [Clinique Breteche, Service de chirurgie plastique, 44 - Nantes (France); Sartre, J Y [Clinique Jules-Verne, Service de chirurgie plastique, 44 - Nantes (France); Malard, O; Ferron, C [CHU Hotel-Dieu, Service de Chirurgie ORL, 44 - Nantes (France); Dreno, B [CHU Hotel-Dieu, Service de dermatologie, 44 - Nantes (France)

    2010-01-15

    Purpose: Evaluation of the results of perioperative interstitial brachytherapy with low dose-rate (L.D.R.) Ir-192 in the treatment of keloid scars. Patients and methods: We performed a retrospective analysis of 73 histologically confirmed keloids (from 58 patients) resistant to medico surgical treated by surgical excision plus early perioperative brachytherapy. All lesions were initially symptomatic. Local control was evaluated by clinical evaluation. Functional and cosmetic results were assessed in terms of patient responses to a self-administered questionnaire. Results: Median age was 28 years (range 13-71 years). Scars were located as follows: 37% on the face, 32% on the trunk or abdomen, 16% on the neck, and 15% on the arms or legs. The mean delay before loading was four hours (range, 1-6 h). The median dose was 20 Gy (range, 15-40 Gy). Sixty-four scars (from 53 patients) were evaluated. Local control was 86% (follow-up, 44.5 months; range, 14-150 months). All relapses occurred early within 2 years posttreatment. At 20 months, survival without recurrence was significantly lower when treated lengths were more than 6 cm long. The rate was 100% for treated scars below 4.5 cm in length, 95% (95% CI: 55-96) for those 4.5-6 cm long, and 75% (95% CI: 56-88) beyond 6 cm (p = 0.038). Of the 35 scars (28 patients) whose results were reassessed, six remained symptomatic and the esthetic results were considered to be good in 51% (18/35) and average in 37% (13/35) (median follow-up, 70 months; range, 16-181 months). Conclusion: Early perioperative L.D.R. brachytherapy delivering 20 Gy at 5 mm reduced the rate of recurrent keloids resistant to other treatments and gave good functional results. (authors)

  10. Outcomes of Node-positive Breast Cancer Patients Treated With Accelerated Partial Breast Irradiation Via Multicatheter Interstitial Brachytherapy: The Pooled Registry of Multicatheter Interstitial Sites (PROMIS) Experience.

    Science.gov (United States)

    Kamrava, Mitchell; Kuske, Robert R; Anderson, Bethany; Chen, Peter; Hayes, John; Quiet, Coral; Wang, Pin-Chieh; Veruttipong, Darlene; Snyder, Margaret; Demanes, David J

    2018-06-01

    To report outcomes for breast-conserving therapy using adjuvant accelerated partial breast irradiation (APBI) with interstitial multicatheter brachytherapy in node-positive compared with node-negative patients. From 1992 to 2013, 1351 patients (1369 breast cancers) were treated with breast-conserving surgery and adjuvant APBI using interstitial multicatheter brachytherapy. A total of 907 patients (835 node negative, 59 N1a, and 13 N1mic) had >1 year of data available and nodal status information and are the subject of this analysis. Median age (range) was 59 years old (22 to 90 y). T stage was 90% T1 and ER/PR/Her2 was positive in 87%, 71%, and 7%. Mean number of axillary nodes removed was 12 (SD, 6). Cox multivariate analysis for local/regional control was performed using age, nodal stage, ER/PR/Her2 receptor status, tumor size, grade, margin, and adjuvant chemotherapy/antiestrogen therapy. The mean (SD) follow-up was 7.5 years (4.6). The 5-year actuarial local control (95% confidence interval) in node-negative versus node-positive patients was 96.3% (94.5-97.5) versus 95.8% (87.6-98.6) (P=0.62). The 5-year actuarial regional control in node-negative versus node-positive patients was 98.5% (97.3-99.2) versus 96.7% (87.4-99.2) (P=0.33). The 5-year actuarial freedom from distant metastasis and cause-specific survival were significantly lower in node-positive versus node-negative patients at 92.3% (82.4-96.7) versus 97.8% (96.3-98.7) (P=0.006) and 91.3% (80.2-96.3) versus 98.7% (97.3-99.3) (P=0.0001). Overall survival was not significantly different. On multivariate analysis age 50 years and below, Her2 positive, positive margin status, and not receiving chemotherapy or antiestrogen therapy were associated with a higher risk of local/regional recurrence. Patients who have had an axillary lymph node dissection and limited node-positive disease may be candidates for treatment with APBI. Further research is ultimately needed to better define specific criteria for APBI

  11. SU-F-T-09: In Phantom Full-Implant Validation of Plastic Scintillation Detectors for in Vivo Dosimetry During Low Dose Rate Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Therriault-Proulx, F; Bruno, T; Beddar, S [UT MD Anderson Cancer Center, Houston, TX (United States); Beaulieu, L [CHU de Quebec, Quebec, QC, CA (Canada)

    2016-06-15

    Purpose: To validate in a water phantom the use of plastic scintillation detectors to measure dose to the urethra and the rectal wall during a clinically realistic low dose rate (LDR) brachytherapy implant. Methods: A template was designed to replicate a clinically realistic LDR brachytherapy prostate implant inside a water phantom. Twenty-two catheters were inserted, including one mimicking the urethra and another the rectal wall. The needles inserted in the remaining 20 catheters were composed of thin-walled nylon tubes in which I-125 radioactive seeds (Air Kerma Strengths of (0.328±0.020)U) were abutted together with plastic spacers to replicate a typical loading. A plastic scintillation detector (PSD) with a 5-mm long × 1-mm diameter sensitive element was first placed inside the urethra and 1-second measurements were performed for 60s after each needle implant. Measurements were also performed at multiple positions along the urethra once all the needles were inserted. The procedure was then repeated with the PSD placed at the rectal wall. Results: Individual dose-rates ranging from 0.07µGy/s to 1.5µGy/s were measured after each needle implant. The average absolute relative differences were (6.2±3.6)% and (6.9±6.5)% to the values calculated with the TG-43 formalism, for the urethra and rectal wall respectively. These results are within expectations from the error uncertainty budget once accounting for uncertainties in seeds’ strength and positioning. Interestingly, the PSD allowed for unplanned error detection as the study was performed. Finally, the measured dose after the full implant at different positions along the mimicked organs at risk were in agreement with TG-43 values for all of the positions tested. Conclusion: Plastic scintillation detectors could be used as in vivo detectors for LDR brachytherapy as they would provide accurate dose information after each needle implant as well as along the organs at risk at the end of the implant.

  12. Brachytherapy needle deflection evaluation and correction

    International Nuclear Information System (INIS)

    Wan Gang; Wei Zhouping; Gardi, Lori; Downey, Donal B.; Fenster, Aaron

    2005-01-01

    In prostate brachytherapy, an 18-gauge needle is used to implant radioactive seeds. This thin needle can be deflected from the preplanned trajectory in the prostate, potentially resulting in a suboptimum dose pattern and at times requiring repeated needle insertion to achieve optimal dosimetry. In this paper, we report on the evaluation of brachytherapy needle deflection and bending in test phantoms and two approaches to overcome the problem. First we tested the relationship between needle deflection and insertion depth as well as whether needle bending occurred. Targeting accuracy was tested by inserting a brachytherapy needle to target 16 points in chicken tissue phantoms. By implanting dummy seeds into chicken tissue phantoms under 3D ultrasound guidance, the overall accuracy of seed implantation was determined. We evaluated methods to overcome brachytherapy needle deflection with three different insertion methods: constant orientation, constant rotation, and orientation reversal at half of the insertion depth. Our results showed that needle deflection is linear with needle insertion depth, and that no noticeable bending occurs with needle insertion into the tissue and agar phantoms. A 3D principal component analysis was performed to obtain the population distribution of needle tip and seed position relative to the target positions. Our results showed that with the constant orientation insertion method, the mean needle targeting error was 2.8 mm and the mean seed implantation error was 2.9 mm. Using the constant rotation and orientation reversal at half insertion depth methods, the deflection error was reduced. The mean needle targeting errors were 0.8 and 1.2 mm for the constant rotation and orientation reversal methods, respectively, and the seed implantation errors were 0.9 and 1.5 mm for constant rotation insertion and orientation reversal methods, respectively

  13. In vivo dosimetry of high-dose-rate brachytherapy: Study on 61 head-and-neck cancer patients using radiophotoluminescence glass dosimeter

    International Nuclear Information System (INIS)

    Nose, Takayuki; Koizumi, Masahiko; Yoshida, Ken; Nishiyama, Kinji; Sasaki, Junichi; Ohnishi, Takeshi; Peiffert, Didier

    2005-01-01

    Purpose: The largest in vivo dosimetry study for interstitial brachytherapy yet examined was performed using new radiophotoluminescence glass dosimeters (RPLGDs). Based on the results, a dose prescription technique achieving high reproducibility and eliminating large hyperdose sleeves was studied. Methods and materials: For 61 head-and-neck cancer patients who underwent high-dose-rate interstitial brachytherapy, new RPLGDs were used for an in vivo study. The Paris System was used for implant. An arbitrary isodose surface was selected for dose prescription. Locations of 83 dosimeters were categorized as on target (n = 52) or on nontarget organ (n = 31) and were also scaled according to % basal dose isodose surface (% BDIS). Compatibility (measured dose/calculated dose) was analyzed according to location. The hyperdose sleeve was assessed in terms of prescription surface expressed in % BDIS. Results: The spread of compatibilities was larger for on nontarget organ (1.06 ± 0.32) than for on target (0.87 ± 0.17, p = 0.01). Within on target RPLGDs, compatibility on 77% and < 95% BDIS for reproducibility and elimination of excessive hyperdose sleeve. For organs at risk, radioprotection should be considered even when calculated dose seems sufficiently low. Further development of planning software is necessary to prevent overestimation

  14. Time factors in breast carcinoma; Influence of delay between external irradiation and brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Dubray, B; Thames, H D [Anderson (M.D.) Hospital and Tumor Inst., Houston, TX (United States); Mazeron, J J; Simon, J M; Pechoux, C le; Calitchi, E; Otmezguine, Y; Bourgeois, J.P. le; Pierquin, B [Hopital Henri-Mondor, 94 - Creteil (France)

    1992-12-01

    From 1971-1983, 398 biopsy-proven breast adenocarcinomas (33 T[sub 1], 309 T[sub 2], 56 T[sub 3]) were treated conservatively at Hopital Henri Mondor by an initial course of external irradiation (45 Gy, 25 fractions, 5 weeks) followed by interstitial iridium-192 implant for a further 37 Gy to the tumor. The mean interval between external irradiation and brachytherapy was 5.9 weeks (S.D. 1.7, range 1-18). 77 local failures were observed at 10-148 months (median 34.5). The actuarial probabilities (S.E.) of local control at 5 and 10 years were 0.86 (0.02) and 0.74 (0.03) respectively. The follow-up for patients free of local recurrence was 4-205 months (median 95). Multivariate analysis showed an increasing probability of local failure with longer interval between external irradiation and brachytherapy (Relative Risk [RR] 1.23 [95% confidence limits: 1.07, 1.41] per week, p=0.005), and a lower risk of failure in case of complete tumor regression after external irradiation (RR 0.47 [0.25, 0.90], p=0.22), and higher brachytherapy dose rate (RR 0.13 [0.02 , 1.02] per Gy/h, p=0.053). No influence of tumor size and total dose (possibly because only limited variations in total dose were observed), or histological grading (not performed in 140 [35%] patients) was found. Because of the lack of dose-control relationship, quantification of the effects of delay between external irradiation and brachytherapy (in terms of compensatory dose) and of dose rate (Incomplete Repair Model) was not possible. The present analysis suggests that the implant dose rate should be high (but no extrapolation can be made above 1 Gy/h) or that total dose should be increased to compensate lower dose rate. However, data do not indicate how much extra dose is necessary, since no dose-control relationship was elicited. In addition, the treatment duration should remain as short as possible in order to maximize local control. (author). 23 refs., 2 figs., 3 tabs.

  15. High dose rate versus low dose rate interstitial radiotherapy for carcinoma of the floor of mouth

    International Nuclear Information System (INIS)

    Inoue, Takehiro; Inoue, Toshihiko; Yamazaki, Hideya; Koizumi, Masahiko; Kagawa, Kazufumi; Yoshida, Ken; Shiomi, Hiroya; Imai, Atsushi; Shimizutani, Kimishige; Tanaka, Eichii; Nose, Takayuki; Teshima, Teruki; Furukawa, Souhei; Fuchihata, Hajime

    1998-01-01

    Purpose: Patients with cancer of the floor of mouth are treated with radiation because of functional and cosmetic reasons. We evaluate the treatment results of high dose rate (HDR) and low dose rate (LDR) interstitial radiation for cancer of the floor of mouth. Methods and Materials: From January 1980 through March 1996, 41 patients with cancer of the floor of mouth were treated with LDR interstitial radiation using 198 Au grains, and from April 1992 through March 1996 16 patients with HDR interstitial radiation. There were 26 T1 tumors, 30 T2 tumors, and 1 T3 tumor. For 21 patients treated with interstitial radiation alone, a total radiation dose of interstitial therapy was 60 Gy/10 fractions/6-7 days in HDR and 85 Gy within 1 week in LDR. For 36 patients treated with a combination therapy, a total dose of 30 to 40 Gy of external radiation and a total dose of 48 Gy/8 fractions/5-6 days in HDR or 65 Gy within 1 week in LDR were delivered. Results: Two- and 5-year local control rates of patients treated with HDR interstitial radiation were 94% and 94%, and those with LDR were 75% and 69%, respectively. Local control rate of patients treated with HDR brachytherapy was slightly higher than that with 198 Au grains (p = 0.113). For late complication, bone exposure or ulcer occurred in 6 of 16 (38%) patients treated with HDR and 13 of 41 (32%) patients treated with LDR. Conclusion: HDR fractionated interstitial brachytherapy can be an alternative to LDR brachytherapy for cancer of the floor of mouth and eliminate radiation exposure for the medical staff

  16. Methods for prostate stabilization during transperineal LDR brachytherapy.

    Science.gov (United States)

    Podder, Tarun; Sherman, Jason; Rubens, Deborah; Messing, Edward; Strang, John; Ng, Wan-Sing; Yu, Yan

    2008-03-21

    In traditional prostate brachytherapy procedures for a low-dose-rate (LDR) radiation seed implant, stabilizing needles are first inserted to provide some rigidity and support to the prostate. Ideally this will provide better seed placement and an overall improved treatment. However, there is much speculation regarding the effectiveness of using regular brachytherapy needles as stabilizers. In this study, we explored the efficacy of two types of needle geometries (regular brachytherapy needle and hooked needle) and several clinically feasible configurations of the stabilization needles. To understand and assess the prostate movement during seed implantation, we collected in vivo data from patients during actual brachytherapy procedures. In vitro experimentation with tissue-equivalent phantoms allowed us to further understand the mechanics behind prostate stabilization. We observed superior stabilization with the hooked needles compared to the regular brachytherapy needles (more than 40% in bilateral parallel needle configuration). Prostate movement was also reduced significantly when regular brachytherapy needles were in an angulated configuration as compared to the parallel configuration (more than 60%). When the hooked needles were angulated for stabilization, further reduction in prostate displacement was observed. In general, for convenience of dosimetric planning and to avoid needle collision, all needles are desired to be in a parallel configuration. In this configuration, hooked needles provide improved stabilization of the prostate. On the other hand, both regular and hooked needles appear to be equally effective in reducing prostate movement when they are in angulated configurations, which will be useful in seed implantation using a robotic system. We have developed nonlinear spring-damper model for the prostate movement which can be used for adapting dosimetric planning during brachytherapy as well as for developing more realistic haptic devices and

  17. Methods for prostate stabilization during transperineal LDR brachytherapy

    International Nuclear Information System (INIS)

    Podder, Tarun; Yu Yan; Sherman, Jason; Rubens, Deborah; Strang, John; Messing, Edward; Ng, Wan-Sing

    2008-01-01

    In traditional prostate brachytherapy procedures for a low-dose-rate (LDR) radiation seed implant, stabilizing needles are first inserted to provide some rigidity and support to the prostate. Ideally this will provide better seed placement and an overall improved treatment. However, there is much speculation regarding the effectiveness of using regular brachytherapy needles as stabilizers. In this study, we explored the efficacy of two types of needle geometries (regular brachytherapy needle and hooked needle) and several clinically feasible configurations of the stabilization needles. To understand and assess the prostate movement during seed implantation, we collected in vivo data from patients during actual brachytherapy procedures. In vitro experimentation with tissue-equivalent phantoms allowed us to further understand the mechanics behind prostate stabilization. We observed superior stabilization with the hooked needles compared to the regular brachytherapy needles (more than 40% in bilateral parallel needle configuration). Prostate movement was also reduced significantly when regular brachytherapy needles were in an angulated configuration as compared to the parallel configuration (more than 60%). When the hooked needles were angulated for stabilization, further reduction in prostate displacement was observed. In general, for convenience of dosimetric planning and to avoid needle collision, all needles are desired to be in a parallel configuration. In this configuration, hooked needles provide improved stabilization of the prostate. On the other hand, both regular and hooked needles appear to be equally effective in reducing prostate movement when they are in angulated configurations, which will be useful in seed implantation using a robotic system. We have developed nonlinear spring-damper model for the prostate movement which can be used for adapting dosimetric planning during brachytherapy as well as for developing more realistic haptic devices and

  18. A robotic device for MRI-guided prostate brachytherapy

    NARCIS (Netherlands)

    Lagerburg, V.

    2008-01-01

    One of the treatment options for prostate cancer is brachytherapy with iodine-125 sources. In prostate brachytherapy a high radiation dose is delivered to the prostate with a steep dose fall off to critical surrounding organs. The implantation of the iodine sources is currently performed under

  19. High dose rate interstitial brachytherapy with external beam irradiation for localized prostate cancer. Preliminary results

    Energy Technology Data Exchange (ETDEWEB)

    Hiratsuka, Junichi; Jo, Yoshimasa; Yoden, Eisaku; Tanaka, Hiroyoshi; Imajo, Yoshinari [Kawasaki Medical School, Kurashiki, Okayama (Japan); Nagase, Naomi; Narihiro, Naomasa; Kubota, Juichi

    2000-12-01

    This study was undertaken to assess the biochemical and pathological results of combined external beam radiotherapy and high dose rate Ir-192 brachytherapy (HDR-Ir192) for clinically localized prostate cancer. Between October 1997 and August 1999, 39 evaluable patients with adenocarcinoma of prostate diagnosed by biopsy were treated with interstitial and external beam irradiation. Patients ranged in age from 58-82 years, with a mean of 69.7 years. T1c, T2 and T3 tumors, according to the UICC classification system (1997), were found in 7, 21 and 11 cases respectively. The mean initial pre-treatment PSA was 35.9 ng/ml (median 13.2), with 77% of the patients having had a pre-treatment PSA greater than 10 ng/ml. Of all patients, 17 had received pre-treatment hormonal therapy. Hormonal pretreatment was stopped at the beginning of radiotherapy in all cases. External beam four-field box irradiation was given to the small pelvis to a dose of 45 Gy/25 fractions. Three HDR-Ir192 treatments were given over a 30-h period, with 5.5 Gy per fraction at the circumference of the prostate gland over the course of this study. Biochemical failure was defined as a PSA level >1.5 ng/ml and rising on three consecutive values. If serial post-treatment PSA levels showed a continuous downward trend, failure was not scored. The patient with clinical evidence of progression was classified as a clinical failure. The median follow-up at the time of evaluation was 19.6 months. A post-treatment PSA level {<=}1.0 ng/ml was seen in 26 (67%) patients, and values from >1.0 to {<=}2.0 ng/ml were seen in 10 (26%) patients. Biochemical failure was not seen in 38 patients except for one patient who developed a distant bone metastasis with negative prostatic biopsy 15 months after treatment. Biochemical control rate was 100% (38/38) except for the patient with bone metastasis classified as clinical failure. Negative biopsies 18 months after treatment were found in 93% (14/15) of patients. Only one patient

  20. Permanent interstitial low-dose-rate brachytherapy for patients with low risk prostate cancer. An interim analysis of 312 cases

    Energy Technology Data Exchange (ETDEWEB)

    Badakhshi, Harun; Graf, Reinhold; Budach, Volker; Wust, Peter [University Hospital Berlin, Department for Radiation Oncology of Charite School of Medicine, Berlin (Germany)

    2015-04-01

    The biochemical relapse-free survival (bRFS) rate after treatment with permanent iodine-125 seed implantation (PSI) or combined seeds and external beam radiotherapy (COMB) for clinical stage T1-T2 localized prostate cancer is a clinically relevant endpoint. The goal of this work was to evaluate the influence of relevant patient- and treatment-related factors. The study population comprised 312 consecutive patients treated with permanent seed implantation. All patients were evaluable for analysis of overall survival (OS) and disease-specific survival (DSS), 230 for bRFS, of which 192 were in the PSI group and 38 in the COMB group. The prescribed minimum peripheral dose was 145 Gy for PSI, for COMB 110 Gy implant and external beam radiotherapy of 45 Gy. The median follow-up time was 33 months (range 8-66 months). bRFS was defined as a serum prostate-specific antigen (PSA) level ≤ 0.2 ng/ml at last follow-up. Overall, the actuarial bRFS at 50 months was 88.4 %. The 50-month bRFS rate for PSI and COMB was 90.9 %, and 77.2 %, respectively. In the univariate analysis, age in the categories ≤ 63 and > 63 years (p < 0.00), PSA nadir (≤ 0.5 ng/ml and > 0.5 ng/ml) and PSA bounce (yes/no) were the significant predicting factors for bRFS. None of the other patient and treatment variables (treatment modality, stage, PSA, Gleason score, risk group, number of risk factors, D90 and various other dose parameters) were found to be a statistically significant predictor of 50-month bRFS. The biochemical failure rates were low in this study. As a proof of principle, our large monocenteric analysis shows that low-dose-rate brachytherapy is an effective and safe procedure for patients with early stage prostate cancer. (orig.) [German] Das biochemisch rezidivfreie Ueberleben (bRFS) nach der Brachytherapie mit permanenter Iod-125-Seed-Implantation (PSI) oder in Kombination mit externer Radiotherapie (COMB) ist beim Patienten mit fruehem Prostatakarzinom (T1/T2) ein relevanter

  1. Early therapy monitoring of 125I seed interstitial implant in a pancreatic cancer xenograft by 18F-FDG Micro-PET/CT

    International Nuclear Information System (INIS)

    Wang Zhongmin; Liu Yu; Chen Kemin; Lu Jian; Gong Ju; Zheng Yunfeng; Zhang Liyun; Liu Fenju

    2011-01-01

    Objective: To investigate the application value of early evaluation and monitoring of 125 I interstitial implantation in a pancreatic cancer xenograft. Methods: Xenograft models were created by subcutaneous injection of Sw 1990 human pancreatic cancer cell suspensions into the right hind limbs of the immunodeficient BABL/c nude mice. The tumors size were about 8-10 mm after two weeks. The mice were randomly divided into 3 groups,including control group (n=4), empty seed implantation group (n=4) and 125 I implantation group (n=4). Before treatment and one week after treatment, 18 F-FDG Micro-PET/CT scan was performed and then maximum standardized uptake values (SUV max ), mean standardized uptake values (SUV mean ), tumor size and necrosis rate were measured. HE staining and TK1 immunohistochemistry examination were carried out in the paraffin-embedded sample. Results: Before treatment the SUV max and SUV mean values of three groups did not reach statistical significance. One week after treatment the SUV max and SUV mean values of three groups were 3.53±1.20 and 0.57±0.26 vs. 3.83±2.13 and 0.59 ±0.24 vs. 0.29±0.23 and 0.016±0.001, respectively, with a significant difference (F=7.62, P=0.01; F=10.34, P=0.005). The SUV max and SUV mean values of 125 I implant group were significantly lower than empty seed implant group and control group and were significantly lower than before treatment. Before treatment, tumor necrosis rate of three groups were not significantly different. Immunohistochemical staining found the TK1 positive staining index of three groups were respectively (64.25±1.71)%, (62.25±2.22)% and (38.25±1.71)% with statistically significant difference (F=233.67, P<0.001). The TK1 positive staining index of 125 I implant group was significantly lower than empty seed implant group and control group. The SUV max values had some positive correlation with TK1 positive staining index (r=0.85, P=0.001). Conclusions: 18 F-FDG Micro-PET/CT may be useful as a

  2. Brachytherapy of carcinoma of vulva with 252Cf

    International Nuclear Information System (INIS)

    Spikalovas, V.; Sinkevicius, V.; Drulia, E.; Kurtinaitis, J.

    1996-01-01

    Thirty patients with carcinoma of vulva were treated with interstitial neutron radiotherapy with 252 Cf. Age of patients was from 32 to 83 years. Stage I was in I patient, stage II - in 12, stage III was in 10 patients. The diagnosis of vulvar cancer was made for the first time in 11 cases, 19 patients had recurrences after the initial treatment. Most of these patients also received external irradiation for cancer of vulva and bilateral inguinal sites with a single fraction dose of 2 Gy to a total dose 30-50 Gy. Enlarged inguinal lymph nodes were irradiated additionally to 60 Gy with reduced field of irradiation. When radiotherapy was used repeatedly interstitial brachytherapy comprised the major part of irradiation dose or the therapy was used alone delivering 35-55 iGy. We used 252 Cf sources with increased activity at the ends 20-30 mm long. A number of inserted sources varied from 2 to 10, irradiation dose rate from 20.3 to 236.7 cGy/h, time of irradiation from 10.2 to 12. hours, RBE from 4.6 to 6.33. Special template device made it possible to implant sources in strictly pre-set geometry. Analysis of survival of patients showed that 2 years survival was 66%, 3 years - 60%, and 5 years survival was 49%. In two cases necrotic epithelitis developed with following radiation ulcer which were cured in 3-4 months. Clinical data showed great effectiveness of interstitial neutron therapy having in mind that 19 patients were treated for recurrences of vulvar cancer after previuos treatment

  3. Modeling of long-range migration of boron interstitials

    International Nuclear Information System (INIS)

    Velichko, O.I.; Burunova, O.N.

    2009-01-01

    A model of the interstitial migration of ion-implanted dopant in silicon during low-temperature thermal treatment has been formulated. It is supposed that the boron interstitials are created during ion implantation or at the initial stage of annealing. During thermal treatment a migration of these impurity interstitials to the surface and in the bulk of a semiconductor occurs. On this basis, a simulation of boron redistribution during thermal annealing for 35 minutes at a temperature of 800 0 C has been carried out. The calculated boron profile agrees well with the experimental data. A number of the parameters describing the interstitial diffusion have been derived. In particular, the average migration length of nonequilibrium boron interstitials is equal to 0.092 μm at a temperature of 800 0 C. To carry out modeling of ion-implanted boron redistribution, the analytical solutions of nonstationary diffusion equation for impurity interstitials have been obtained. The case of Dirichlet boundary conditions and the case of reflecting boundary on the surface of a semiconductor have been considered. (authors)

  4. Computed tomography in brachytherapy

    International Nuclear Information System (INIS)

    Mansfield, C.M.; Lee, K.R.; Dwyer, S.; Zellmer, D.; Cook, P.

    1983-01-01

    CT scanning adds to the ability to evaluate brachytherapy techniques. It provides an additional method in the assessment of patients who are candidates for or who are being treated by brachytherapy. The CT scan can give information regarding the position of the sources and their relation to the tumor and normal structures with greater ease than do orthogonal views. This makes it possible to accurately calculate areas of high or low dose. Potential areas of overdose can be recognized, thereby decreasing the chances of postbrachytherapy complications. CT scanning can be used at various levels of complexity in dosimetry evaluation. Adequate brachytherapy dosimetry information is obtainable from CT slices through one or more levels of the implanted volume. In some instances it is possible to obtain additional information by reconstructing the scans in other planes, e.g., coronal or sagittal. Three-dimensional viewing of the implant is desirable, but it should be pointed out that this approach is time-consuming and beyond the capabilities of most institutions at present. It will be necessary to continue work on three-dimensional treatment planning to make it readily available

  5. Impact of pre-implant lower urinary tract symptoms on postoperative urinary morbidity after permanent prostate brachytherapy

    International Nuclear Information System (INIS)

    Teishima, Jun; Iwamoto, Hideo; Miyamoto, Katsutoshi; Shoji, Koichi; Masumoto, Hiroshi; Inoue, Shogo; Kobayashi, Kanao; Kajiwara, Mitsuru; Matsubara, Akio

    2012-01-01

    prolongation. The present findings suggest that the presence of lower urinary tract symptoms before implantation does not prolong urinary morbidity after permanent prostate brachytherapy. (author)

  6. Cosmetic Analysis Following Breast-Conserving Surgery and Adjuvant High-Dose-Rate Interstitial Brachytherapy for Early-Stage Breast Cancer: A Prospective Clinical Study

    International Nuclear Information System (INIS)

    Garsa, Adam A.; Ferraro, Daniel J.; DeWees, Todd; Margenthaler, Julie A.; Naughton, Michael; Aft, Rebecca; Gillanders, William E.; Eberlein, Timothy; Matesa, Melissa A.; Zoberi, Imran

    2013-01-01

    Purpose: To prospectively evaluate cosmetic outcomes in women treated with accelerated partial breast irradiation using high-dose-rate interstitial brachytherapy for early-stage breast cancer. Methods and Materials: Between 2004 and 2008, 151 patients with early-stage breast cancer were enrolled in a phase 2 prospective clinical trial. Eligible patients had stage Tis-T2 tumors of ≤3 cm that were excised with negative margins and with no nodal involvement. Patients received 3.4 Gy twice daily to a total dose of 34 Gy. Both the patients and the treating radiation oncologist qualitatively rated cosmesis as excellent, good, fair, or poor over time and ascribed a cause for changes in cosmesis. Cosmetic outcome was evaluated quantitatively by percentage of breast retraction assessment (pBRA). Patients also reported their satisfaction with treatment over time. Results: Median follow-up was 55 months. The rates of excellent-to-good cosmesis reported by patients and the treating radiation oncologist were 92% and 97% pretreatment, 91% and 97% at 3 to 4 months' follow-up, 87% and 94% at 2 years, and 92% and 94% at 3 years, respectively. Breast infection and adjuvant chemotherapy were independent predictors of a fair-to-poor cosmetic outcome at 3 years. Compared to pretreatment pBRA (7.35), there was no significant change in pBRA over time. The volume receiving more than 150 Gy (V150) was the only significant predictor of pBRA. The majority of patients (86.6%) were completely satisfied with their treatment. Conclusions: Patients and the treating physician reported a high rate of excellent-to-good cosmetic outcomes at all follow-up time points. Acute breast infection and chemotherapy were associated with worse cosmetic outcomes. Multicatheter interstitial brachytherapy does not significantly change breast size as measured by pBRA

  7. Cosmetic Analysis Following Breast-Conserving Surgery and Adjuvant High-Dose-Rate Interstitial Brachytherapy for Early-Stage Breast Cancer: A Prospective Clinical Study

    Energy Technology Data Exchange (ETDEWEB)

    Garsa, Adam A.; Ferraro, Daniel J.; DeWees, Todd [Department of Radiation Oncology, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Margenthaler, Julie A. [Department of Surgery, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Naughton, Michael [Department of Medicine, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Aft, Rebecca [Department of Surgery, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Department of Surgery, John Cochran Veterans Hospital, St. Louis, Missouri (United States); Gillanders, William E.; Eberlein, Timothy [Department of Surgery, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Matesa, Melissa A. [Department of Radiation Oncology, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Zoberi, Imran, E-mail: izoberi@radonc.wustl.edu [Department of Radiation Oncology, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States)

    2013-03-15

    Purpose: To prospectively evaluate cosmetic outcomes in women treated with accelerated partial breast irradiation using high-dose-rate interstitial brachytherapy for early-stage breast cancer. Methods and Materials: Between 2004 and 2008, 151 patients with early-stage breast cancer were enrolled in a phase 2 prospective clinical trial. Eligible patients had stage Tis-T2 tumors of ≤3 cm that were excised with negative margins and with no nodal involvement. Patients received 3.4 Gy twice daily to a total dose of 34 Gy. Both the patients and the treating radiation oncologist qualitatively rated cosmesis as excellent, good, fair, or poor over time and ascribed a cause for changes in cosmesis. Cosmetic outcome was evaluated quantitatively by percentage of breast retraction assessment (pBRA). Patients also reported their satisfaction with treatment over time. Results: Median follow-up was 55 months. The rates of excellent-to-good cosmesis reported by patients and the treating radiation oncologist were 92% and 97% pretreatment, 91% and 97% at 3 to 4 months' follow-up, 87% and 94% at 2 years, and 92% and 94% at 3 years, respectively. Breast infection and adjuvant chemotherapy were independent predictors of a fair-to-poor cosmetic outcome at 3 years. Compared to pretreatment pBRA (7.35), there was no significant change in pBRA over time. The volume receiving more than 150 Gy (V150) was the only significant predictor of pBRA. The majority of patients (86.6%) were completely satisfied with their treatment. Conclusions: Patients and the treating physician reported a high rate of excellent-to-good cosmetic outcomes at all follow-up time points. Acute breast infection and chemotherapy were associated with worse cosmetic outcomes. Multicatheter interstitial brachytherapy does not significantly change breast size as measured by pBRA.

  8. Limited external irradiation and interstitial 192iridium implant in the treatment of squamous cell carcinoma of the tonsillar region

    International Nuclear Information System (INIS)

    Puthawala, A.A.; Syed, A.M.; Eads, D.L.; Neblett, D.; Gillin, L.; Gates, T.C.

    1985-01-01

    Between January 1976 and March 1982, 80 patients with histologically proven diagnosis of squamous cell carcinoma of the tonsillar region were treated with definitive radiotherapy. Sixty-five (81%) of these patients had locally advanced tumors (Stage III and IV); 49% of patients had clinically palpable cervical lymphadenopathy. All patients received a combined external megavoltage and interstitial irradiation. The dose of external irradiation was limited to 4500-5000 cGy over 41/2 to 51/2 weeks. This was followed by interstitial 192 iridium implants to doses of 2000-2500 cGy in 50-60 hours for T1, T2 lesions and 3000-4000 cGy in 60-100 hours for T3, T4 lesions. The neck masses were also separately implanted to deliver additional doses of 2000-4000 cGy in 50-80 hours. Overall local tumor control was observed in 84% of patients with a minimum follow-up period of 2 years. An absolute 3-year disease free survival of the entire group was 72%. Treatment related complications such as soft tissue necrosis or osteoradionecrosis occurred in 6% (5/80) of patients. The salvage of neck failures and local failures was possible in 78 and 38% of patients, respectively, either by surgery or by re-irradiation employing interstitial 192 iridium implants. Functional and esthetic integrity was well preserved in most cases

  9. Survey of brachytherapy practice in the United States: a report of the Clinical Research Committee of the American Endocurietherapy Society.

    Science.gov (United States)

    Nag, S; Owen, J B; Farnan, N; Pajak, T F; Martinez, A; Porter, A; Blasko, J; Harrison, L B

    1995-01-01

    To obtain reliable data on the extent of the brachytherapy practice in the United States by conducting a comprehensive survey of all facilities. The Clinical Research Committee of the AES surveyed all 1321 radiation oncology facilities identified in the Patterns of Care Study (PCS) of the American College of Radiology (ACR). Multiple mailings and follow-up were made to obtain a high response rate. Survey responders and nonresponders were compared using chi-square tests. Summary statistics were reported. Of the 1321 facilities, 1054 responded (80%). Hospital-based and larger facilities had a statistically significant higher rate of response. Brachytherapy was being performed at 819 facilities (the median number of procedures = 21-50). Two hundred and two facilities did no brachytherapy. The common isotopes used were 137Cs (705 facilities), 192Ir (585 facilities), 125I (236 facilities), and 131I (194 facilities). The common brachytherapy techniques used were intracavitary (751 facilities), interstitial (536 facilities), intraluminal (310 facilities), and plaques (148 facilities). Remote afterloaded brachytherapy was used at 205 centers as follows: high dose rate (HDR) (164), medium dose rate (MDR) (5), and low dose rate (LDR) (36). Computerized dosimetry was most commonly used (790 facilities), followed by Patterson-Parker (104 facilities) and Quimby (72 facilities). The common sites treated were cervix (701 facilities), endometrium (565 facilities), head and neck (354 facilities), and lung (344 facilities). Data regarding brachytherapy practice has been obtained from a large percentage (80%) of all facilities in the United States. The majority (78-81%) of radiation oncology facilities perform brachytherapy; however, its use is restricted to gynecological implants in many of these centers. The results from this survey will be used to develop a pattern of care study and data registry in brachytherapy.

  10. Experimental determination of dosimetric characterization of a newly designed encapsulated interstitial brachytherapy source of 103Pd-model Pd-1

    International Nuclear Information System (INIS)

    Nath, Ravinder; Yue Ning; Roa, Eduardo

    2002-01-01

    A newly designed encapsulated 103 Pd source has been introduced (BrachySeed trade mark sign -Pd-103, also named Model Pd-1, manufactured by DRAXIMAGE Inc. and distributed by Cytogen Corp.) for interstitial brachytherapy to provide more isotropic dose distributions. In this work, the dosimetric characteristics of the 103 Pd source were measured with micro LiF TLD chips and dosimetry parameters were characterized based upon the American Association of Physicists in Medicine (AAPM) Task Group No. 43 formalism. The dose rate constant of the sources was determined to be 0.66±0.05 cGy h-1 U-1. The radial dose function was measured and was found to be similar to that of the Theragenics Model 200 103 Pd source. The anisotropy constant for the Model Pd-1 source was determined to be 1.03

  11. High dose rate brachytherapy for oral cancer.

    Science.gov (United States)

    Yamazaki, Hideya; Yoshida, Ken; Yoshioka, Yasuo; Shimizutani, Kimishige; Furukawa, Souhei; Koizumi, Masahiko; Ogawa, Kazuhiko

    2013-01-01

    Brachytherapy results in better dose distribution compared with other treatments because of steep dose reduction in the surrounding normal tissues. Excellent local control rates and acceptable side effects have been demonstrated with brachytherapy as a sole treatment modality, a postoperative method, and a method of reirradiation. Low-dose-rate (LDR) brachytherapy has been employed worldwide for its superior outcome. With the advent of technology, high-dose-rate (HDR) brachytherapy has enabled health care providers to avoid radiation exposure. This therapy has been used for treating many types of cancer such as gynecological cancer, breast cancer, and prostate cancer. However, LDR and pulsed-dose-rate interstitial brachytherapies have been mainstays for head and neck cancer. HDR brachytherapy has not become widely used in the radiotherapy community for treating head and neck cancer because of lack of experience and biological concerns. On the other hand, because HDR brachytherapy is less time-consuming, treatment can occasionally be administered on an outpatient basis. For the convenience and safety of patients and medical staff, HDR brachytherapy should be explored. To enhance the role of this therapy in treatment of head and neck lesions, we have reviewed its outcomes with oral cancer, including Phase I/II to Phase III studies, evaluating this technique in terms of safety and efficacy. In particular, our studies have shown that superficial tumors can be treated using a non-invasive mold technique on an outpatient basis without adverse reactions. The next generation of image-guided brachytherapy using HDR has been discussed. In conclusion, although concrete evidence is yet to be produced with a sophisticated study in a reproducible manner, HDR brachytherapy remains an important option for treatment of oral cancer.

  12. Demonstration of brachytherapy boost dose-response relationships in glioblastoma multiforme

    International Nuclear Information System (INIS)

    Sneed, Penny K.; Lamborn, Kathleen R.; Larson, David A.; Prados, Michael D.; Malec, Mary K.; McDermott, Michael W.; Weaver, Keith A.; Phillips, Theodore L.; Wara, William M.; Gutin, Philip H.

    1996-01-01

    Purpose: To evaluate brachytherapy dose-response relationships in adults with glioblastoma undergoing temporary 125 I implant boost after external beam radiotherapy. Methods and Materials: Since June 1987, orthogonal radiographs using a fiducial marker box have been used to verify brain implant source positions and generate dose-volume histograms at the University of California, San Francisco. For adults who underwent brachytherapy boost for glioblastoma from June 1987 through December 1992, tumor volumes were reoutlined to ensure consistency and dose-volume histograms were recalculated. Univariate and multivariate analyses of various patient and treatment parameters were performed evaluating for influence of dose on freedom from local failure (FFLF) and actuarial survival. Results: Of 102 implant boosts, 5 were excluded because computer plans were unavailable. For the remaining 97 patients, analyses with adjustment for known prognostic factors (age, KPS, extent of initial surgical resection) and prognostic factors identified on univariate testing (adjuvant chemotherapy) showed that higher minimum brachytherapy tumor dose was strongly associated with improved FFLF (p = 0.001). A quadratic relationship was found between total biological effective dose and survival, with a trend toward optimal survival probability at 47 Gy minimum brachytherapy tumor dose (corresponding to about 65 Gy to 95% of the tumor volume); survival decreased with lower or higher doses. Two patients expired and one requires hospice care because of brain necrosis after brachytherapy doses > 63 Gy to 95% of the tumor volume with 60 Gy to > 18 cm 3 of normal brain. Conclusion: Although higher minimum brachytherapy tumor dose was strongly associated with better local control, a brachytherapy boost dose > 50-60 Gy may result in life-threatening necrosis. We recommend careful conformation of the prescription isodose line to the contrast enhancing tumor volume, delivery of a minimum brachytherapy

  13. Implant volume as a prognostic variable in brachytherapy decision-making for malignant gliomas stratified by the RTOG recursive partitioning analysis

    International Nuclear Information System (INIS)

    Videtic, Gregory M.M.; Gaspar, Laurie E.; Zamorano, Lucia; Stitt, Larry W.; Fontanesi, James; Levin, Kenneth J.

    2001-01-01

    Purpose: When an initial retrospective review of malignant glioma patients (MG) undergoing brachytherapy was carried out using the Radiation Therapy Oncology Group (RTOG) recursive partitioning analysis (RPA) criteria, it revealed that glioblastoma multiforme (GBM) cases benefit the most from implant. In the present study, we focused exclusively on these GBM patients stratified by RPA survival class and looked at the relationship between survival and implanted target volume, to distinguish the prognostic value of volume in general and for a given GBM class. Methods and Materials: Between 1991 and 1998, 75 MG patients were treated with surgery, external beam radiation, and stereotactic iodine-125 (I-125) implant. Of these, 53 patients (70.7%) had GBMs, with 52 (98%) having target volume (TV) data for analysis. Stratification by RPA criteria showed 12, 26, 13, and 1 patients in classes III to VI, respectively. For analysis purposes, classes V and VI were merged. There were 27 (51.9%) male and 25 (48.1%) female patients. Mean age was 57.5 years (range 14-79). Median Karnofsky performance status (KPS) was 90 (range 50-100). Median follow-up time was 11 months (range 2-79). Results: At analysis, 18 GBM patients (34.6%) were alive and 34 (65.4%) were dead. Two-year and 5-year survivals were 42% and 17.5%, respectively, with a median survival time (MST) of 16 months. Two-year survivals and MSTs for the implanted GBM patients compared to the RTOG database were as follows: 74% vs. 35% and 28 months vs. 17.9 months for class III; 32% vs. 15% and 16 months vs. 11.1 months for class IV; 29% vs. 6% and 11 months vs. 8.9 months for class V/VI. Mean implanted TV was 15.5 cc (range 0.8-78), which corresponds to a spherical implant diameter of 3.1 cm. Plotting survival as a function of 5-cc TV increments suggested a trend toward poorer survival as the implanted volume increases. The impact of incremental changes in TV on survival within a given RPA class of GBMs was compared to the

  14. Effects of seed migration on post-implant dosimetry of prostate brachytherapy

    International Nuclear Information System (INIS)

    Gao, M.; Wang, J. Z.; Nag, S.; Gupta, N.

    2007-01-01

    Brachytherapy using permanent seed implants has been an effective treatment for prostate cancer. However, seeds will migrate after implant, thus making the evaluation of post-implant dosimetry difficult. In this study, we developed a computer program to simulate seed migration and analyzed dosimetric changes due to seed migration at various migration amounts. The study was based on 14 patients treated with Pd-103 at the James Cancer Hospital. Modeling of seed migration, including direction, distance as well as day of migration, was based on clinical observations. Changes of commonly used dosimetric parameters as a function of migration amount (2, 4, 6 mm respectively), prostate size (from 20 to 90 cc), and prostate region (central vs peripheral) were studied. Change of biological outcome (tumor control probability) due to migration was also estimated. Migration reduced prostate D90 to 99±2% of original value in 2 mm migration, and the reduction increased to 94±6% in 6 mm migration. The reduction of prostate dose led to a 14% (40%) drop in the tumor control probability for 2 mm (6 mm) migration, assuming radiosensitive tumors. However, migration has less effect on a prostate implanted with a larger number of seeds. Prostate V100 was less sensitive to migration than D90 since its mean value was still 99% of original value even in 6 mm migration. Migration also showed a different effect in the peripheral region vs the central region of the prostate, where the peripheral mean dose tended to drop more significantly. Therefore, extra activity implanted in the peripheral region during pre-plan can be considered. The detrimental effects of migration were more severe in terms of increasing the dose to normal structures, as rectum V50 may be 70% higher and urethra V100 may be 50% higher in the case of 6 mm migration. Quantitative knowledge of these effects is helpful in treatment planning and post-implant evaluation

  15. The Adjoint Method for The Optimization of Brachytherapy and Radiotherapy Patient Treatment Planning Procedures Using Monte Carlo Calculations

    International Nuclear Information System (INIS)

    Henderson, D.L.; Yoo, S.; Kowalok, M.; Mackie, T.R.; Thomadsen, B.R.

    2001-01-01

    The goal of this project is to investigate the use of the adjoint method, commonly used in the reactor physics community, for the optimization of radiation therapy patient treatment plans. Two different types of radiation therapy are being examined, interstitial brachytherapy and radiotherapy. In brachytherapy radioactive sources are surgically implanted within the diseased organ such as the prostate to treat the cancerous tissue. With radiotherapy, the x-ray source is usually located at a distance of about 1-meter from the patient and focused on the treatment area. For brachytherapy the optimization phase of the treatment plan consists of determining the optimal placement of the radioactive sources, which delivers the prescribed dose to the disease tissue while simultaneously sparing (reducing) the dose to sensitive tissue and organs. For external beam radiation therapy the optimization phase of the treatment plan consists of determining the optimal direction and intensity of beam, which provides complete coverage of the tumor region with the prescribed dose while simultaneously avoiding sensitive tissue areas. For both therapy methods, the optimal treatment plan is one in which the diseased tissue has been treated with the prescribed dose and dose to the sensitive tissue and organs has been kept to a minimum

  16. CT-guided interstitial 125I seed implantation for intractable pelvic recurrence of rectal carcinomas

    International Nuclear Information System (INIS)

    Lin Zhenwen; Chu Hong; Kong Jian; Dou Yongchong

    2011-01-01

    Objective: To evaluate the therapeutic effect and safety of CT-guided interstitial 125 I seed implantation in treating intractable pelvic recurrence of rectal carcinomas. Methods: During the period from April 2010 to September 2010 CT-guided interstitial 125 I seed implantation was carried out in 11 patients with pelvic recurrence of rectal carcinoma which developed after the patients had received comprehensive treatments such as surgical resection, chemotherapy and/or radiotherapy. The clinical data were retrospectively analyzed. The clinical symptoms, the tumor size and the quality of life (QOL) before the treatment and at two and six months after the treatment were recorded, and the complications were observed. The results were compared. All the patients were followed up for six months. Results: At two and six months after the treatment, the improvement rate of the clinical symptoms was 100.0% (11/11) and 90.9% (10/11) respectively, while the effective reduction of the tumor size was 90.9% (10/11) and 81.8% (9/11) respectively. At two and six months after the treatment the QOL score was (56.0±3.66) and (54.4±5.41) respectively, both of which were higher than the QOL score determined before the treatment (42.5±6.93), the difference between them was statistically significant (P 125 I seed implantation has satisfactory short-term effect. This treatment is also quite safe and effective for patients who are unbearable to external radiation therapy due to the tissue dose restriction. (authors)

  17. A Prospective Longitudinal Clinical Trial Evaluating Quality of Life After Breast-Conserving Surgery and High-Dose-Rate Interstitial Brachytherapy for Early-Stage Breast Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Garsa, Adam A.; Ferraro, Daniel J.; DeWees, Todd A. [Department of Radiation Oncology, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Deshields, Teresa L. [Department of Medicine, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Margenthaler, Julie A.; Cyr, Amy E. [Department of Surgery, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Naughton, Michael [Department of Medicine, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Aft, Rebecca [Department of Surgery, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Department of Surgery, John Cochran Veterans Hospital, St. Louis, Missouri (United States); Gillanders, William E.; Eberlein, Timothy [Department of Surgery, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Matesa, Melissa A.; Ochoa, Laura L. [Department of Radiation Oncology, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States); Zoberi, Imran, E-mail: izoberi@radonc.wustl.edu [Department of Radiation Oncology, Siteman Cancer Center, Barnes-Jewish Hospital and Washington University School of Medicine, St. Louis, Missouri (United States)

    2013-12-01

    Purpose: To prospectively examine quality of life (QOL) of patients with early stage breast cancer treated with accelerated partial breast irradiation (APBI) using high-dose-rate (HDR) interstitial brachytherapy. Methods and Materials: Between March 2004 and December 2008, 151 patients with early stage breast cancer were enrolled in a phase 2 prospective clinical trial. Eligible patients included those with Tis-T2 tumors measuring ≤3 cm excised with negative surgical margins and with no nodal involvement. Patients received 3.4 Gy twice daily to a total dose of 34 Gy. QOL was measured using European Organization for Research and Treatment of Cancer (EORTC) QLQ-C30, version 3.0, and QLQ-BR23 questionnaires. The QLQ-C30 and QLQ-BR23 questionnaires were evaluated during pretreatment and then at 6 to 8 weeks, 3 to 4 months, 6 to 8 months, and 1 and 2 years after treatment. Results: The median follow-up was 55 months. Breast symptom scores remained stable in the months after treatment, and they significantly improved 6 to 8 months after treatment. Scores for emotional functioning, social functioning, and future perspective showed significant improvement 2 years after treatment. Symptomatic fat necrosis was associated with several changes in QOL, including increased pain, breast symptoms, systemic treatment side effects, dyspnea, and fatigue, as well as decreased role functioning, emotional functioning, and social functioning. Conclusions: HDR multicatheter interstitial brachytherapy was well tolerated, with no significant detrimental effect on measured QOL scales/items through 2 years of follow-up. Compared to pretreatment scores, there was improvement in breast symptoms, emotional functioning, social functioning, and future perspective 2 years after treatment.

  18. A Prospective Longitudinal Clinical Trial Evaluating Quality of Life After Breast-Conserving Surgery and High-Dose-Rate Interstitial Brachytherapy for Early-Stage Breast Cancer

    International Nuclear Information System (INIS)

    Garsa, Adam A.; Ferraro, Daniel J.; DeWees, Todd A.; Deshields, Teresa L.; Margenthaler, Julie A.; Cyr, Amy E.; Naughton, Michael; Aft, Rebecca; Gillanders, William E.; Eberlein, Timothy; Matesa, Melissa A.; Ochoa, Laura L.; Zoberi, Imran

    2013-01-01

    Purpose: To prospectively examine quality of life (QOL) of patients with early stage breast cancer treated with accelerated partial breast irradiation (APBI) using high-dose-rate (HDR) interstitial brachytherapy. Methods and Materials: Between March 2004 and December 2008, 151 patients with early stage breast cancer were enrolled in a phase 2 prospective clinical trial. Eligible patients included those with Tis-T2 tumors measuring ≤3 cm excised with negative surgical margins and with no nodal involvement. Patients received 3.4 Gy twice daily to a total dose of 34 Gy. QOL was measured using European Organization for Research and Treatment of Cancer (EORTC) QLQ-C30, version 3.0, and QLQ-BR23 questionnaires. The QLQ-C30 and QLQ-BR23 questionnaires were evaluated during pretreatment and then at 6 to 8 weeks, 3 to 4 months, 6 to 8 months, and 1 and 2 years after treatment. Results: The median follow-up was 55 months. Breast symptom scores remained stable in the months after treatment, and they significantly improved 6 to 8 months after treatment. Scores for emotional functioning, social functioning, and future perspective showed significant improvement 2 years after treatment. Symptomatic fat necrosis was associated with several changes in QOL, including increased pain, breast symptoms, systemic treatment side effects, dyspnea, and fatigue, as well as decreased role functioning, emotional functioning, and social functioning. Conclusions: HDR multicatheter interstitial brachytherapy was well tolerated, with no significant detrimental effect on measured QOL scales/items through 2 years of follow-up. Compared to pretreatment scores, there was improvement in breast symptoms, emotional functioning, social functioning, and future perspective 2 years after treatment

  19. Pulsed-dose-rate peri-operative brachytherapy as an interstitial boost in organ-sparing treatment of breast cancer

    Directory of Open Access Journals (Sweden)

    Krystyna Serkies

    2016-12-01

    Full Text Available Purpose : To evaluate peri-operative multicatheter interstitial pulsed-dose-rate brachytherapy (PDR-BT with an intra-operative catheter placement to boost the tumor excision site in breast cancer patients treated conservatively. Material and methods: Between May 2002 and October 2008, 96 consecutive T1-3N0-2M0 breast cancer patients underwent breast-conserving therapy (BCT including peri-operative PDR-BT boost, followed by whole breast external beam radiotherapy (WBRT. The BT dose of 15 Gy (1 Gy/pulse/h was given on the following day after surgery. Results: No increased bleeding or delayed wound healing related to the implants were observed. The only side effects included one case of temporary peri-operative breast infection and 3 cases of fat necrosis, both early and late. In 11 patients (11.4%, subsequent WBRT was omitted owing to the final pathology findings. These included eight patients who underwent mastectomy due to multiple adverse prognostic pathological features, one case of lobular carcinoma in situ, and two cases with no malignant tumor. With a median follow-up of 12 years (range: 7-14 years, among 85 patients who completed BCT, there was one ipsilateral breast tumor and one locoregional nodal recurrence. Six patients developed distant metastases and one was diagnosed with angiosarcoma within irradiated breast. The actuarial 5- and 10-year disease free survival was 90% (95% CI: 84-96% and 87% (95% CI: 80-94%, respectively, for the patients with invasive breast cancer, and 91% (95% CI: 84-97% and 89% (95% CI: 82-96%, respectively, for patients who completed BCT. Good cosmetic outcome by self-assessment was achieved in 58 out of 64 (91% evaluable patients. Conclusions : Peri-operative PDR-BT boost with intra-operative tube placement followed by EBRT is feasible and devoid of considerable toxicity, and provides excellent long-term local control. However, this strategy necessitates careful patient selection and histological confirmation

  20. High dose rate brachytherapy for oral cancer

    International Nuclear Information System (INIS)

    Yamazaki, Hideya; Yoshida, Ken; Yoshioka, Yasuo; Shimizutani, Kimishige; Koizumi, Masahiko; Ogawa, Kazuhiko; Furukawa, Souhei

    2013-01-01

    Brachytherapy results in better dose distribution compared with other treatments because of steep dose reduction in the surrounding normal tissues. Excellent local control rates and acceptable side effects have been demonstrated with brachytherapy as a sole treatment modality, a postoperative method, and a method of reirradiation. Low-dose-rate (LDR) brachytherapy has been employed worldwide for its superior outcome. With the advent of technology, high-dose-rate (HDR) brachytherapy has enabled health care providers to avoid radiation exposure. This therapy has been used for treating many types of cancer such as gynecological cancer, breast cancer, and prostate cancer. However, LDR and pulsed-dose-rate interstitial brachytherapies have been mainstays for head and neck cancer. HDR brachytherapy has not become widely used in the radiotherapy community for treating head and neck cancer because of lack of experience and biological concerns. On the other hand, because HDR brachytherapy is less time-consuming, treatment can occasionally be administered on an outpatient basis. For the convenience and safety of patients and medical staff, HDR brachytherapy should be explored. To enhance the role of this therapy in treatment of head and neck lesions, we have reviewed its outcomes with oral cancer, including Phase I/II to Phase III studies, evaluating this technique in terms of safety and efficacy. In particular, our studies have shown that superficial tumors can be treated using a non-invasive mold technique on an outpatient basis without adverse reactions. The next generation of image-guided brachytherapy using HDR has been discussed. In conclusion, although concrete evidence is yet to be produced with a sophisticated study in a reproducible manner, HDR brachytherapy remains an important option for treatment of oral cancer. (author)

  1. Correlations of post-implant regional dosimetric parameters at 24 hours and one month, with clinical results of low-dose-rate brachytherapy for localized prostate cancer

    Directory of Open Access Journals (Sweden)

    Eiichiro Okazaki

    2017-12-01

    Full Text Available Purpose : To evaluate the correlations of post-implant regional dosimetrics at 24 hours (24 h and 1 month after implant procedures, with clinical outcomes of low-dose-rate (LDR brachytherapy for localized prostate cancer. Material and methods : Between January 2008 and December 2014, 130 consecutive patients treated for localized prostate cancer, receiving definitive iodine-125 ( 125 I brachytherapy treatment were retrospectively analyzed. All patients underwent post-implant CT imaging for dosimetric analysis at 24 h and 1 month after implantation procedure. Prostate contours were divided into quadrants: anterior-superior (ASQ, posterior-superior (PSQ, anterior-inferior (AIQ, and posterior-inferior (PIQ. Predictive factors and cut-off values of biochemical failure-free survival (BFFS and toxicities of LDR brachytherapy were analyzed. Results : The median follow-up time was 69.5 months. Seven patients (5.4% had biochemical failure. The 3-year and 5-year BFFS rates were 96.7% and 93.1%, respectively. On multivariate analysis, prostate-specific antigen and Gleason score were significant prognostic factors for biochemical failure. D 90 (the minimal dose received by 90% of the volume of PSQ and PIQ at 24 h, and D 90 of PSQ at 1 month were also significant factors. The cut-off values of PSQ D 90 were 145 Gy at 24 h and 160 Gy at 1 month. D 90 of the whole prostate was not significant at 24 h and at 1 month. D 90 of PSQ at 1 month was a significant factor for rectal hemorrhage. Conclusions : Post-implant D 90 of PSQ is significantly associated with BFFS for localized prostate cancer not only at 1 month, but also at 24 hours. D 90 of PSQ at 1 month is also a significant factor for rectal hemorrhage.

  2. Intracranial interstitial radiation

    International Nuclear Information System (INIS)

    Willis, D.; Rittenmeyer, H.; Hitchon, P.

    1986-01-01

    Primary malignant brain tumors are fatal, with 90% of patients having these tumors dying within two years following diagnosis. Cranial interstitial radiation therapy, a technique under investigation to control these tumors, involves implantation of radioactive iodine 125 seeds into the tumor bed by stereotaxic technique. The interstitial radiation technique, monitoring of radiation, and nursing care of patients are discussed. Case histories are presented, along with discussion of results attained using this therapy, and its future

  3. Poster - Thur Eve - 03: LDR to HDR: RADPOS applications in brachytherapy.

    Science.gov (United States)

    Cherpak, A J; Cygler, J E; Kertzscher, G; E, C; Perry, G

    2012-07-01

    The RADPOS in vivo dosimetry system combines an electromagnetic positioning sensor and either one or five MOSFET dosimeters. The feasibility of using the system for quality control has been explored for a range of radiotherapy treatment techniques including most recently transperineal interstitial permanent prostate brachytherapy and high dose rate (HDR) treatments. Dose and position information was collected by a RADPOS array detector inside a Foley catheter within patients' urethra during permanent seed implantation. Ten patients were studied, and average displacement during implantation was Δr = (1.4-5.1) mm, with movements up to 9.7 mm due to the removal of the transrectal ultrasound probe. Maximum integral dose in the prostatic urethra ranged from 110-195 Gy, and it was found that the dose can change up to 63 cGy (62.0%) depending on whether the rectal probe is in place. For HDR, a RADPOS detector was first calibrated with an Ir-192 source. A treatment was then simulated using a total of 50 dwell positions in 5 catheters in an acrylic phantom. Dwell positions ranged from 1 to 10 cm away from the RADPOS detector and dose was measured for each source position. An average calibration coefficient of 0.74±0.11 cGy/mV was calculated for the detector and the average absolute difference between measured values and expected dose was 0.7±5.4 cGy (5±20%). The demonstrated accuracy of RADPOS dose measurements along with its ability to simultaneously measure displacement makes it a powerful tool for brachytherapy treatments, where high dose gradients can present unique in vivo dosimetry challenges. © 2012 American Association of Physicists in Medicine.

  4. A magazine for the implantation of I-125 seeds in interstitial therapy of cerebral tumors

    International Nuclear Information System (INIS)

    Bockermann, V.; Dieckmann, G.; Lott, H.

    1988-01-01

    A compact magazine has been developed allowing a quick and easy implantation of I-125 seeds in interstitial therapy of intracranial tumors. The magazine is made of steel and has the shape of a right parallelepiped. One seed can be put each time into a movable insert. The magazine is connected to a normal application canula. It is easy to handle and contributes much to radioprotection. (orig.) [de

  5. The usefulness of metal markers for CTV-based dose prescription in high-dose-rate interstitial brachytherapy

    International Nuclear Information System (INIS)

    Yoshida, Ken; Mitomo, Masanori; Nose, Takayuki; Koizumi, Masahiko; Nishiyama, Kinji; Yoshida, Mineo

    2002-01-01

    We employ a clinical target volume (CTV)-based dose prescription for high-dose-rate (HDR) interstitial brachytherapy. However, it is not easy to define CTV and organs at risk (OAR) from X-ray film or CT scanning. To solve this problem, we have utilized metal markers since October 1999. Moreover, metal markers can help modify dose prescription. By regulating the doses to the metal markers, refining the dose prescription can easily be achieved. In this research, we investigated the usefulness of the metal markers. Between October 1999 and May 2001, 51 patients were implanted with metal markers at Osaka Medical Center for Cancer and Cardiovascular Diseases (OMCC), Osaka National Hospital (ONH) and Sanda City Hospital (SCH). Forty-nine patients (head and neck: 32; pelvis: 11; soft tissue: 3; breast: 3) using metal markers were analyzed. During operation, we implanted 179 metal markers (49 patients) to CTV and 151 markers (26 patients) to OAR. At treatment planning, CTV was reconstructed judging from the metal markers, applicator position and operation records. Generally, we prescribed the tumoricidal dose to an isodose surface that covers CTV. We also planned to limit the doses to OAR lower than certain levels. The maximum normal tissue doses were decided 80%, 150%, 100%, 50% and 200% of the prescribed doses for the rectum, the urethra, the mandible, the skin and the large vessel, respectively. The doses to the metal markers using CTV-based dose prescription were generated. These were compared with the doses theoretically calculated with the Paris system. Treatment results were also investigated. The doses to the 158 metal markers (42 patients) for CTV were higher than ''tumoricidal dose''. In 7 patients, as a result of compromised dose prescription, 9 markers were lower than the tumoricidal dose. The other 12 markers (7%) were excluded from dose evaluation because they were judged as miss-implanted. The doses to the 142 metal markers (24 patients) for OAR were lower

  6. Efficacy of prophylactic single-dose therapy using fluoroquinolone for prostate brachytherapy

    International Nuclear Information System (INIS)

    Nomura, Takeo; Hirai, Kenichi; Yamasaki, Mutsushi; Inoue, Toru; Takahashi, Mika; Kawashima, Takayuki; Sato, Fuminori; Mimata, Hiromitsu

    2012-01-01

    There is little definitive evidence to guide the use of prophylactic antibiotics for prostate brachytherapy. The purpose of this study is to evaluate the incidence of postimplant infections in patients who receive antimicrobial prophylaxis with pazufloxacin (PZFX). A total of 84 patients who underwent prostate brachytherapy received a single intravenous dose of PZFX at 500 mg perioperatively for 1 day. No postimplant antibiotic medication was prescribed. Urinalysis, plasma white blood cell (WBC) count, and C reactive protein (CRP) levels were evaluated before the implantation, on the day after implantation, and on the 7th and 28th days after brachytherapy. None of the 84 patients (0.0%) developed a symptomatic urinary tract infection or had febrile infectious complications after brachytherapy. There were statistically significant elevations in the levels of erythrocytes, leukocytes, bacteria in urine, plasma WBC and CRP postoperatively, but these values did not exceed the normal range or were only slightly elevated on the day after brachytherapy (day 1) and on day 7. All laboratory examinations had returned to the normal range on day 28. Single-dose therapy with fluoroquinolone helps to prevent infections after prostate brachytherapy. (author)

  7. The use of high-dose-rate brachytherapy alone after lumpectomy in patients with early-stage breast cancer treated with breast-conserving therapy

    International Nuclear Information System (INIS)

    Baglan, Kathy L.; Martinez, Alvaro A.; Frazier, Robert C.; Kini, Vijay R.; Kestin, Larry L.; Chen, Peter Y.; Edmundson, Greg; Mele, Elizabeth; Jaffray, David; Vicini, Frank A.

    2001-01-01

    Purpose: We present the preliminary results of our in-house protocol using outpatient high-dose-rate (HDR) brachytherapy as the sole radiation modality following lumpectomy in patients with early-stage breast cancer. Methods and Materials: Thirty-seven patients with 38 Stage I-II breast cancers received radiation to the lumpectomy cavity alone using an HDR interstitial implant with 192 Ir. A minimum dose of 32 Gy was delivered on an outpatient basis in 8 fractions of 4 Gy to the lumpectomy cavity plus a 1- to 2-cm margin over consecutive 4 days. Results: Median follow-up is 31 months. There has been one ipsilateral breast recurrence for a crude failure rate of 2.6% and no regional or distant failures. Wound healing was not impaired in patients undergoing an open-cavity implant. Three minor breast infections occurred, and all resolved with oral antibiotics. The cosmetic outcome was good to excellent in all patients. Conclusion: In selected patients with early-stage breast cancer, treatment of the lumpectomy cavity alone with outpatient HDR brachytherapy is both technically feasible and well tolerated. Early results are encouraging, however, longer follow-up is necessary before equivalence to standard whole-breast irradiation can be established and to determine the most optimal radiation therapy technique to be employed

  8. The needs for brachytherapy source calibrations in the United States

    International Nuclear Information System (INIS)

    Coursey, B.M.; Goodman, L.J.; Hoppes, D.D.; Loevinger, R.; McLaughlin, W.L.; Soares, C.G.; Weaver, J.T.

    1992-01-01

    Brachytherapy sources of beta and gamma radiation ('brachy' is from the Greek, meaning 'near') have a long history of use in interstitial, intracavitary, intraluminal, and ocular radiation therapy. In the past the US national standards for these sources were often specified in activity or milligram radium equivalent. With the introduction of new radionuclide sources to replace radium, source strength calibrations are now expressed as air kerma rate at a meter. In this paper, we review the NIST standards for brachytherapy sources, list some of the common radionuclides and source encapsulations in use in the US radiology community, and describe the latest NIST work, in collaboration with several US medical institutions, on a method of two- and three-dimensional dose mapping of brachytherapy sources using radiochromic films. (orig.)

  9. Three-dimensional verification of 125I seed stability after permanent implantation in the parotid gland and periparotid region

    International Nuclear Information System (INIS)

    Fan, Yi; Huang, Ming-Wei; Zheng, Lei; Zhao, Yi-Jiao; Zhang, Jian-Guo

    2015-01-01

    To evaluate seed stability after permanent implantation in the parotid gland and periparotid region via a three-dimensional reconstruction of CT data. Fifteen patients treated from June 2008 to June 2012 at Peking University School and Hospital of Stomatology for parotid gland tumors with postoperative adjunctive 125 I interstitial brachytherapy were retrospectively reviewed in this study. Serial CT data were obtained during follow-up. Mimics and Geomagic Studio software were used for seed reconstruction and stability analysis, respectively. Seed loss and/or migration outside of the treated area were absent in all patients during follow-up (23–71 months). Total seed cluster volume was maximized on day 1 post-implantation due to edema and decreased significantly by an average of 13.5 % (SD = 9.80 %; 95 % CI, 6.82–17.68 %) during the first two months and an average of 4.5 % (SD = 3.60 %; 95 % CI, 2.29–6.29 %) during the next four months. Volume stabilized over the subsequent six months. 125 I seed number and location were stable with a general volumetric shrinkage tendency in the parotid gland and periparotid region. Three-dimensional seed reconstruction of CT images is feasible for visualization and verification of implanted seeds in parotid brachytherapy

  10. Predictive factors for acute and late urinary toxicity after permanent interstitial brachytherapy in Japanese patients

    International Nuclear Information System (INIS)

    Tanimoto, Ryuta; Bekku, Kensuke; Katayama, Norihisa

    2013-01-01

    The objectives of this study were to describe the frequency of and to determine predictive factors associated with Radiation Therapy Oncology Group urinary toxicity in prostate brachytherapy patients. From January 2004 to April 2011, 466 consecutive Japanese patients underwent permanent iodine-125-seed brachytherapy (median follow up 48 months). International Prostate Symptom Score and Radiation Therapy Oncology Group toxicity data were prospectively collected. Prostate volume, International Prostate Symptom Score before and after brachytherapy, and postimplant analysis were examined for an association with urinary toxicity, defined as Radiation Therapy Oncology Group urinary toxicity of Grade 1 or higher. Logistic regression analysis was used to examine the factors associated with urinary toxicity. The rate of Radiation Therapy Oncology Group urinary toxicity grade 1 or higher at 1, 6, 12, 24, 36 and 48 months was 67%, 40%, 21%, 31%, 27% and 28%, respectively. Grade 2 or higher urinary toxicity was less than 1% at each time-point. International Prostate Symptom Score was highest at 3 months and returned to normal 12 months after brachytherapy. On multivariate analysis, patients with a larger prostate size, greater baseline International Prostate Symptom Score, higher prostate V100, higher prostate V150, higher prostate D90 and a greater number of seeds had more acute urinary toxicities at 1 month and 12 months after brachytherapy. On multivariate analysis, significant predictors for urinary toxicity at 1 month and 12 months were a greater baseline International Prostate Symptom Score and prostate V100. Most urinary symptoms are tolerated and resolved within 12 months after prostate brachytherapy. Acute and late urinary toxicity after brachytherapy is strongly related to the baseline International Prostate Symptom Score and prostate V100. (author)

  11. SU-F-T-03: Radiobiological Evaluation of a Directional Brachytherapy Device Surgically Implanted Following EBRT

    Energy Technology Data Exchange (ETDEWEB)

    Rivard, MJ [Tufts University School of Medicine, Boston, MA (United States); Emrich, JG; Poli, J [Drexel University College of Medicine, Philadelphia, PA (United States)

    2016-06-15

    Purpose: Preceding surgical implantation following external-beam radiotherapy (EBRT) delivery, a radiobiological evaluation was performed for a new LDR Pd-103 directional brachytherapy device (CivaSheet). As this was the first case with the device used in combination with EBRT, there was concern to determine the appropriate prescription dose. Methods: The radiobiological model of Dale (1985, 1989) was used for a permanent LDR implant including radioactive decay. The biological effective dose (BED) was converted to the equivalent dose in 2 Gy fractions (EQD2) for comparison with EBRT prescription expectations. Given IMRT delivery of 50.4 Gy, an LDR brachytherapy dose of approximately 15–20 Gy EQD2 was desired. To be specific to the treatment site (leiomyosarcoma T2bN0M0, grade 2 with R1 surgical margin), the radiobiological model required several radiobiological parameters with values taken from the literature. A sensitivity analysis was performed to determine their relative importance on the calculated BED and subsequent EQD2. The Pd-103 decay constant (λ=0.0017 h{sup −1}) was also used. DVHs were prepared for pre- and post-surgical geometries to glean the possible and realized implant geometric configuration. DVHs prepared in VariSeed9 were converted to BEDVHs and subsequently EQD2 values for each volume-element. Results: For a physical dose of 28 Gy to a 0.5 cm depth, BED=21.7 Gy and EQD2=17.6 Gy, which was near the center of the desired EQD2 range. Tumor bed (CTV=4 cm{sup 3}) coverage was 99.2% with 48 sources implanted. In order of decreasing importance from the sensitivity analysis, the radiobiological parameters were α=0.25 Gy{sup −1}, T{sub POT}=23 days, α/β=8.6 Gy, and T=1.5 h. Percentage variations in these values produced EQD2 variations of 40%, 20%, 18%, and 1%, respectively. Conclusion: This radiobiological evaluation indicated that prescription dose may be determined for comparison with the desired EQD2, and that radiobiologicalparameter

  12. Phantom study of radiation doses outside the target volume brachytherapy versus external radiotherapy of early breast cancer

    International Nuclear Information System (INIS)

    Johansson, Bengt; Persson, Essie; Westman, Gunnar; Persliden, Jan

    2003-01-01

    Background and purpose: Brachytherapy is sometimes suggested as an adjuvant treatment after surgery of some tumours. When introducing this, it would be useful to have an estimate of the dose distribution to different body sites, both near and distant to target, comparing conventional external irradiation to brachytherapy. The aim of the present study was to determine radiation doses with both methods at different body sites, near and distant to target, in an experimental situation on an operated left sided breast cancer on a female Alderson phantom. Methods: Five external beam treatments with isocentric tangential fields were given by a linear accelerator. A specified dose of 1.0 Gy was given to the whole left sided breast volume. Five interstitial brachytherapy treatments were given to the upper, lateral quadrant of the left breast by a two plane, 10 needles implant. A dose of 1.0 Gy specified according to the Paris system was administered by a pulsed dose rate afterloading machine. Absorbed dose in different fixed dose points were measured by thermoluminescence dosimeters. Results: Both methods yielded an absorbed dose of the same size to the bone marrow and internal organs distant to target, 1.0-1.4% of the prescribed dose. There was a trend of lower doses to the lower half of the trunk and higher doses to the upper half of the trunk, respectively, by brachytherapy. A 90% reduction of absorbed dose with brachytherapy compared to external irradiation was found in the near-target region within 5 cm from target boundary where parts of the left lung and the heart are situated. If an adjuvant dose of 50 Gy is given with the external radiotherapy and brachytherapy, the absorbed dose in a part of the myocardium could be reduced from 31.8 to 2.1 Gy. Conclusions: Near target, brachytherapy yielded a considerably lower absorbed dose which is of special importance when considering radiation effects on the myocard and lungs. We could not demonstrate any difference of

  13. Treatment results of Stage I and II oral tongue cancer with interstitial brachytherapy: maximum tumor thickness is prognostic of nodal metastasis

    International Nuclear Information System (INIS)

    Matsuura, Kanji; Hirokawa, Yutaka; Fujita, Minoru; Akagi, Yukio; Ito, Katsuhide

    1998-01-01

    Purpose: To evaluate the prognostic importance of T classification and maximum tumor thickness (MTT) on the treatment results of Stage I and II oral tongue cancer treated with interstitial brachytherapy. Methods and Materials: Between January 1981 and December 1993, 173 cases were eligible for this retrospective analysis. Of 173 patients, 75 were classified as Stage I and 98 as Stage II: maximum tumor length ranged from 6 to 40 mm. MTT, which ranged from 2 to 38 mm, was measured with ultrasonography and/or palpation. Brachytherapy was performed with iridium hairpins or radium needles following external irradiation in 66 patients, or exclusively in 107 patients. Results: The 5-year local recurrence rates were Stage I, 7%; Stage II, 22%; MTT < 8 mm, 8%; and MTT ≥ 8 mm, 28%. The 5-year regional recurrence rates were Stage I, 15%; Stage II, 29%; MTT < 8 mm, 18%; and MTT ≥ 8 mm, 31%, respectively. The 5-year local recurrence rates of the patients with Stage I and MTT < 8 mm of the brachytherapy only group were significantly better than those of Stage II and MTT ≥ 8 mm (5% and 6% vs. 16% and 24%). The 5-year regional recurrence rates of the patients with Stage I and MTT < 8 mm of the brachytherapy-only group were significantly better than those of Stage II and MTT ≥ 8 mm (14% and 16% vs. 34% and 46%). There was no significant difference in the 5-year regional recurrence rates between the two groups of Stage I and Stage II, MTT < 8 mm. However, there was a significant difference in the 5-year regional recurrence rates between the two groups of MTT ≥ 8 mm (p < 0.005). Conclusions: For patients with Stage I and II oral tongue cancer, tumor thickness as well as T classification were prognostic for nodal metastasis and prognosis. Patients with MTT ≥ 8 mm are more likely to fail in the neck region. These findings suggest that MTT should be considered along with T stage in determining strategies for Stage I and II oral tongue cancer

  14. Quality of life following 3D conformal radiation therapy or permanent interstitial brachytherapy for localized prostate cancer

    International Nuclear Information System (INIS)

    Michalski, J.M.; Kong, F.M.; Mansur, D.B.; Ahmed, N.; Perez, C.A.

    2001-01-01

    Purpose: Both 3D Conformal Radiation Therapy (3DCRT) and Transperineal Interstitial Permanent Brachytherapy (TIPPB) are offered as suitable non-surgical alternatives to radical prostatectomy. Despite equivalent cancer control, very little data has been published that compares Quality of Life (QOL) in contemporary cohorts of patients choosing these treatments. Materials and Methods: Since 1998, patients selecting either 3DCRT alone or TIPPB (monotherapy or boost after external beam) for primary management of localized prostate cancer were asked to participate in a prospective assessment of QOL measures. In this preliminary report, 41 3DCRT and 40 TIPPB (34 monotherapy, 6 boost) patients completed validated QOL instruments at each followup visit. QOL instruments included the International Prostate Symptom Score (IPSS), FACT-P, and Sexual Adjustment Questionnaire (SAQ). Results: The average age of men in each group was 69 years. Choice of treatment was left to the patient unless there were significant medical or technical contraindications to either modality. 3DCRT total doses ranged from 61-78 Gy (mean 73.5Gy) and TIPPB doses were 145Gy (TG43) in 34 I-125 implants and 115 Gy in 1 Pd-103 (monotherapy) or 90 Gy in 5 Pd-103 (boost) implants. Patients undergoing TIPPB reported significantly worse urinary and sexual function than their counterparts receiving 3DCRT. The mean cumulative IPSS was 12.5 with TIPPB compared to 8.3 with 3DCRT (p=0.036). Differences were most pronounced in the first 12 months after treatment, particularly with respect to the strength of stream and the need to strain. TIPPB patients were more likely to report a need to urinate frequently (p=0.02), require a pad (p=0.001), be bothered (p=0.02), or have activity limited by urinary side effects (p=0.01). TIPPB patients were less likely to resume sexual activity within 6 months after treatment (p=0.0003) and engaged in sexual activity less often (p= 0.016) than 3DCRT patients. They were also more

  15. Verification of the calculation program for brachytherapy planning system of high dose rate (PLATO)

    International Nuclear Information System (INIS)

    Almansa, J.; Alaman, C.; Perez-Alija, J.; Herrero, C.; Real, R. del; Ososrio, J. L.

    2011-01-01

    In our treatments are performed brachytherapy high dose rate since 2007. The procedures performed include gynecological intracavitary treatment and interstitial. The treatments are performed with a source of Ir-192 activity between 5 and 10 Ci such that small variations in treatment times can cause damage to the patient. In addition the Royal Decree 1566/1998 on Quality Criteria in radiotherapy establishes the need to verify the monitor units or treatment time in radiotherapy and brachytherapy. All this justifies the existence of a redundant system for brachytherapy dose calculation that can reveal any abnormality is present.

  16. The application of early post-operative combination therapy with interstitial hyperthermia in the treatment of malignant gliomas

    International Nuclear Information System (INIS)

    Stahl, H.; Wust, P.; Maier-Hauff, K.; Seebass, M.; Mischel, M.; Grammler, M.; Golde, G.; Loeffel, J.; Felix, R.

    1995-01-01

    A pilot study was performed to examine practicality, tolerability, effectiveness and scope for improvement in early postoperative hyperthermia treatment following catheter implantation as part of (partial) tumor resection. Each CT data set was transferred into a VAX 3100 workstation for retrospective analysis of the hyperthermia treatment. The implanted catheters were segmented and the distributions of power density and temperature were simulated. We sought to achieve the best possible temperature distributions by optimising the catheter arrangement in the planning calculations. The corresponding Ir-192-source brachytherapy treatments were simulated in a similar way using the implanted, as well as optimised catheter arrays. Intraoperative catheter implantation in 4 patients was problem-free. Postoperative complications were not observed, neither were infections. Interstitial microwave hyperthermia in combination with percutaneous irradiation or chemotherapy a few days after the operation was also tolerated well by all patients. Effective temperatures (of at least 42 C) were regularly achieved at measurement points, but the temperature distributions were unsatisfactory, with T 90 values (the temperature reached in at least 90% of the target volume) of under 38 C. Measured temperature/position curves showed qualitative correlation with the simulated calculations. The catheter positions determined by optimisation varied significantly from the positions clinically used. (orig./MG) [de

  17. Validation of mathematical models for the prediction of organs-at-risk dosimetric metrics in high-dose-rate gynecologic interstitial brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Damato, Antonio L.; Viswanathan, Akila N.; Cormack, Robert A. [Dana-Farber Cancer Institute and Brigham and Women' s Hospital, Boston, Massachusetts 02115 (United States)

    2013-10-15

    Purpose: Given the complicated nature of an interstitial gynecologic brachytherapy treatment plan, the use of a quantitative tool to evaluate the quality of the achieved metrics compared to clinical practice would be advantageous. For this purpose, predictive mathematical models to predict the D{sub 2cc} of rectum and bladder in interstitial gynecologic brachytherapy are discussed and validated.Methods: Previous plans were used to establish the relationship between D2cc and the overlapping volume of the organ at risk with the targeted area (C0) or a 1-cm expansion of the target area (C1). Three mathematical models were evaluated: D{sub 2cc}=α*C{sub 1}+β (LIN); D{sub 2cc}=α– exp(–β*C{sub 0}) (EXP); and a mixed approach (MIX), where both C{sub 0} and C{sub 1} were inputs of the model. The parameters of the models were optimized on a training set of patient data, and the predictive error of each model (predicted D{sub 2cc}− real D{sub 2cc}) was calculated on a validation set of patient data. The data of 20 patients were used to perform a K-fold cross validation analysis, with K = 2, 4, 6, 8, 10, and 20.Results: MIX was associated with the smallest mean prediction error <6.4% for an 18-patient training set; LIN had an error <8.5%; EXP had an error <8.3%. Best case scenario analysis shows that an error ≤5% can be achieved for a ten-patient training set with MIX, an error ≤7.4% for LIN, and an error ≤6.9% for EXP. The error decreases with the increase in training set size, with the most marked decrease observed for MIX.Conclusions: The MIX model can predict the D{sub 2cc} of the organs at risk with an error lower than 5% with a training set of ten patients or greater. The model can be used in the development of quality assurance tools to identify treatment plans with suboptimal sparing of the organs at risk. It can also be used to improve preplanning and in the development of real-time intraoperative planning tools.

  18. Toxicity and cosmetic result of partial breast high-dose-rate interstitial brachytherapy for conservatively operated early breast cancer

    International Nuclear Information System (INIS)

    Xiu Xia; Tripuraneni Prabhakar; Giap Huan; Lin Ray; Chu Colin

    2007-01-01

    Objective: Objective To study the method, side effects and cosmetic outcome of high- dose-rate (HDR) accelerated partial breast interstitial irradiation (APBI) alone in early stage breast cancer' after conservative surgery. Methods: From February 2002 to June 2003,47 breast cancer lesions from 46 patients suffering from stage I/II breast cancer were treated with HDR 192 Ir APBI after conservative surgery. All patients were over 40 year-old, with T1-2N0-1 (≤3 lymph nodes positive), surgical margin > 1-2 mm, but those having lobular or inflammatory breast cancer were excluded. HDR brachytherapy with 34 Gy, 10 fractions/5 days was used after surgery, toxic reaction and cosmetic outcome were observed in one month, 6 and 12 months respectively. Results: Follow up of 1846 months, 34 months was carried out for the whole group. During the treatment, acute reactions including: erythema, edema, tenderness and infection, all under I-II grade, none of III-IV grade were observed in 21 patients(46%); late toxicity reactions: skin fibrosis, breast tenderness, fat necrosis, and telangiectasia, totally 20 patients (43%) were observed: 2 patients in III grade but one patient received 6 cycle chemotherapy. The result of cosmetic outcome evaluation was excellent or good, at 6 months 95% and 12 months 98%, respectively, but there was no recurfence. Conclusions: Excellent and favorable cosmetic results are noted after APBI by interstitial alone. Acute and late reactions are few. Long term observation is necessary for the rate of' local control. (authors)

  19. SU-E-T-378: Evaluation of An Analytical Model for the Inter-Seed Attenuation Effect in 103-Pd Multi-Seed Implant Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Safigholi, H; Soliman, A; Song, W [Sunnybrook Research Institute, Sunnybrook Health Sciences Centre, U of T, Toronto, Ontario (Canada); Meigooni, A Soleimani [Comprehensive Cancer Center of Nevada, Las Vegas, Nevada (United States); Han, D [Sunnybrook Research Institute, Sunnybrook Health Sciences Centre, U of T, Toronto, Ontario (Canada); University of California, San Diego, La Jolla, CA (United States)

    2015-06-15

    Purpose: Brachytherapy treatment planning systems based on TG-43 protocol calculate the dose in water and neglects the heterogeneity effect of seeds in multi-seed implant brachytherapy. In this research, the accuracy of a novel analytical model that we propose for the inter-seed attenuation effect (ISA) for 103-Pd seed model is evaluated. Methods: In the analytical model, dose perturbation due to the ISA effect for each seed in an LDR multi-seed implant for 103-Pd is calculated by assuming that the seed of interest is active and the other surrounding seeds are inactive. The cumulative dosimetric effect of all seeds is then summed using the superposition principle. The model is based on pre Monte Carlo (MC) simulated 3D kernels of the dose perturbations caused by the ISA effect. The cumulative ISA effect due to multiple surrounding seeds is obtained by a simple multiplication of the individual ISA effect by each seed, the effect of which is determined by the distance from the seed of interest. This novel algorithm is then compared with full MC water-based simulations (FMCW). Results: The results show that the dose perturbation model we propose is in excellent agreement with the FMCW values for a case with three seeds separated by 1 cm. The average difference of the model and the FMCW simulations was less than 8%±2%. Conclusion: Using the proposed novel analytical ISA effect model, one could expedite the corrections due to the ISA dose perturbation effects during permanent seed 103-Pd brachytherapy planning with minimal increase in time since the model is based on multiplications and superposition. This model can be applied, in principle, to any other brachytherapy seeds. Further work is necessary to validate this model on a more complicated geometry as well.

  20. Role of brachytherapy in treatment of epidermoid carcinomas of the vallecula after conservative supraglottic laryngectomy followed by irradiation

    International Nuclear Information System (INIS)

    Clippe, Sebastien; Pommier, Pascal; Poupart, Marc; Ceruse, Philippe; Rosenbusch, Thomas; Ramade, Antoine; Montbarbon, Xavier; Gerard, Jean-Pierre; Carrie, Christian; Ardiet, Jean-Michel

    2002-01-01

    Purpose:To evaluate survival and functional results of the treatment of carcinomas of the vallecula using surgery, irradiation, and interstitial brachytherapy. Methods and Materials: Between 1990 and 1998, 36 patients with squamous cell carcinoma of the vallecula were treated with horizontal supraglottic functional laryngectomy, external beam radiotherapy (median dose 54 Gy), and additional interstitial brachytherapy (median dose 16 Gy). Results were compared with a previous series of 22 patients treated without brachytherapy. Results: The median follow-up was 44 months. The 5-year actuarial overall survival rate was 61.3%. The 5-year specific survival rate was 86%, with 2 local failures (local control rate 94.4%) and 4 isolated distant metastases. Ten patients developed a second primary. The overall survival was 34% for 22 patients previously treated without brachytherapy. Severe toxicities occurred in 9 patients: death (related to larynx edema or inhalation, n=1), soft tissue necrosis (n=1), aspiration pneumonia (n=1), mandibular necrosis (n=2), pharyngocutaneous fistula (n=2), and laryngeal edema (n=2). All the patients fed orally with no definitive gastrostomy or tracheotomy. Conclusion: Additional brachytherapy for vallecula carcinoma seems to improve locoregional control and overall survival dramatically. Functional results were also excellent. To our knowledge, this original therapeutic schedule has never been previously described

  1. SU-E-J-215: Towards MR-Only Image Guided Identification of Calcifications and Brachytherapy Seeds: Application to Prostate and Breast LDR Implant Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Elzibak, A; Fatemi-Ardekani, A; Soliman, A; Mashouf, S; Safigholi, H; Ravi, A; Morton, G; Song, WY [Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario (Canada); Han, D [Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario (Canada); University of California, San Diego, La Jolla, CA (United States)

    2015-06-15

    Purpose: To identify and analyze the appearance of calcifications and brachytherapy seeds on magnitude and phase MRI images and to investigate whether they can be distinguished from each other on corrected phase images for application to prostate and breast low dose rate (LDR) implant dosimetry. Methods: An agar-based gel phantom containing two LDR brachytherapy seeds (Advantage Pd-103, IsoAid, 0.8mm diameter, 4.5mm length) and two spherical calcifications (large: 7mm diameter and small: 4mm diameter) was constructed and imaged on a 3T Philips MR scanner using a 16-channel head coil and a susceptibility weighted imaging (SWI) sequence (2mm slices, 320mm FOV, TR/ TE= 26.5/5.3ms, 15 degree flip angle). The phase images were unwrapped and corrected using a 32×32, 2D Hanning high pass filter to remove background phase noise. Appearance of the seeds and calcifications was assessed visually and quantitatively using Osirix (http://www.osirix-viewer.com/). Results: As expected, calcifications and brachytherapy seeds appeared dark (hypointense) relative to the surrounding gel on the magnitude MRI images. The diameter of each seed without the surrounding artifact was measured to be 0.1 cm on the magnitude image, while diameters of 0.79 and 0.37 cm were measured for the larger and smaller calcifications, respectively. On the corrected phase images, the brachytherapy seeds and the calcifications appeared bright (hyperintense). The diameter of the seeds was larger on the phase images (0.17 cm) likely due to the dipole effect. Conclusion: MRI has the best soft tissue contrast for accurate organ delineation leading to most accurate implant dosimetry. This work demonstrated that phase images can potentially be useful in identifying brachytherapy seeds and calcifications in the prostate and breast due to their bright appearance, which helps in their visualization and quantification for accurate dosimetry using MR-only. Future work includes optimizing phase filters to best identify

  2. Experimental and theoretical determination of dosimetric characteristics of IsoAid ADVANTAGETM125I brachytherapy source

    International Nuclear Information System (INIS)

    Meigooni, Ali S.; Hayes, Joshua L.; Zhang Hualin; Sowards, Keith

    2002-01-01

    125 I brachytherapy sources are being used for interstitial implants in tumor sites such as the prostate. Recently, the ADVANTAGE TM 125 I, Model IAI-125, source became commercially available for interstitial brachytherapy treatment. Dosimetric characteristics (dose rate constant, radial dose function, and anisotropy function) of this source were experimentally and theoretically determined, following the AAPM Task Group 43 recommendations. Derivation of the dose rate constant was based on recent NIST WAFAC calibration performed in accordance with their 1999 standard. Measurements were performed in Solid Water TM phantom using LiF thermoluminescent dosimeters. The theoretical calculations were performed in both Solid Water TM and water using the PTRAN Monte Carlo code. The results indicated that a dose rate constant of the new source in water was 0.98±0.03 cGy h -1 U -1 . The radial dose function of the new source was measured in Solid Water TM and calculated both in water and Solid Water TM at distances up to 10.0 cm. The anisotropy function, F(r,θ), of the new source was measured and calculated in Solid Water TM at distances of 2 cm, 3 cm, 5 cm, and 7 cm and also was calculated in water at distances ranging from 1 cm to 7 cm from the source. From the anisotropy function, the anisotropy factors and anisotropy constant were derived. The anisotropy constant of the ADVANTAGE TM 125 I source in water was found to be 0.97±0.03. The dosimetric characteristics of this new source compared favorably with those from the Amersham Health Model 6711 source. Complete dosimetric parameters of the new source are presented in this paper

  3. Local control after brachytherapy for localized prostatic carcinomas

    International Nuclear Information System (INIS)

    Wachter, T.; Peneau, M.; Sabattier, R.; Breteau, N.

    1996-01-01

    From 1991 to 1995; 31 patients (mid-age: 70 years) underwent prostatic brachytherapy for localized prostate cancers using Iridium 192 transperineal percutaneous interstitial implantation guided by transrectal ultrasonography. Initial staging included among other evaluations a bilateral staging, iliac and obturator lymph nodes dissection. Classification according to stage was : T1b=16%, T1c=36%, T2a=19%, T2b=13%, T2c=13%, T3a=3%. All patients were N (-). Gleason score was 5 for 55%. 77% of the initial PSA was < 25μg/l. Follow-up included one clinical control and psa determination at 1-3-6-12 and 18 months, bone scanning at 12 months and prostate biopsy guided by transrectal ultrasonography at 18, 24, 30 months. Up to now, mean follow-up is 32 months. At one month, psa was normal (< 2,5μg/l) in 21% of the patients, at 12 months 60% and 67% two years after brachytherapy. Biopsies at 18 months were negative for 60% of the patients and 63% at 24 months. 3 patients were metastased after 3 years. 4 patients had severe complications with colostomy and/or urinary derivation. This technic seems to be interesting for localized prostate cancers T1 and T2 with initial psa < 25μg/l. Two third of the patients had normal psa and negative biopsies after 2 years. The rate of ano-rectal and urinary morbidity is high but is explained by the technic used at the beginning of this study

  4. Brachytherapy for T1-T2 floor-of-the-mouth cancers: the Gustave-Roussy Institute experience

    International Nuclear Information System (INIS)

    Marsiglia, Hugo; Haie-Meder, Christine; Sasso, Giuseppe; Mamelle, Gerard; Gerbaulet, Alain

    2002-01-01

    Purpose: In a retrospective analysis, we evaluated the Gustave-Roussy Institute's experience of locoregional control, survival, and complications of low-dose rate brachytherapy for carcinoma of the floor of the mouth. Methods and Materials: Between 1970 and 1985, 160 patients with previously untreated carcinoma of the floor of the mouth received interstitial brachytherapy as definitive treatment. Of the 160 patients, 79 (49%) had T1 and 81 (51%) had T2 lesions, and 127 (79%) had N0 and 33 (21%) had N1; 84% of tumors arose from the anterior floor of the mouth. Brachytherapy was performed with 192 Ir wires, according to the Paris system rules, followed by neck dissection (T2 or N1) or follow-up (T1N0). Results: With a follow-up period of 9-19 years, the observed survival rates were 89% at 2 years and 76% at 5 years, and the local control rates were 93% in T1 and 88% in T2 tumors. A low rate of distant metastases was noticed (5%); 31% of patients developed a second primary cancer. Severe mucosal necrosis was observed in <10% of patients. Any grade of bone necrosis was seen in 18% of cases (only 2.5% had G3 necrosis). This complication occurred more frequently in patients with poor dental status and in those treated without dental protection during implantation (p <0.001). Conclusion: Radical brachytherapy offers excellent local control (89%) and an acceptable rate of complications (<10% severe necrosis) that may be significantly decreased with dental care and the use of protective devices. The high incidence of second malignancies remains a major concern in these patients

  5. Seed Placement in Permanent Breast Seed Implant Brachytherapy: Are Concerns Over Accuracy Valid?

    Energy Technology Data Exchange (ETDEWEB)

    Morton, Daniel, E-mail: dmorton@bccancer.bc.ca [Department of Medical Physics, BC Cancer Agency, Centre for the Southern Interior, Kelowna, British Columbia (Canada); Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia (Canada); Hilts, Michelle [Department of Medical Physics, BC Cancer Agency, Centre for the Southern Interior, Kelowna, British Columbia (Canada); Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia (Canada); Batchelar, Deidre [Department of Medical Physics, BC Cancer Agency, Centre for the Southern Interior, Kelowna, British Columbia (Canada); Crook, Juanita [Department of Radiation Oncology, BC Cancer Agency, Centre for the Southern Interior, Kelowna, British Columbia (Canada)

    2016-07-01

    Purpose: To evaluate seed placement accuracy in permanent breast seed implant brachytherapy (PBSI), to identify any systematic errors and evaluate their effect on dosimetry. Methods and Materials: Treatment plans and postimplant computed tomography scans for 20 PBSI patients were spatially registered and used to evaluate differences between planned and implanted seed positions, termed seed displacements. For each patient, the mean total and directional seed displacements were determined in both standard room coordinates and in needle coordinates relative to needle insertion angle. Seeds were labeled according to their proximity to the anatomy within the breast, to evaluate the influence of anatomic regions on seed placement. Dosimetry within an evaluative target volume (seroma + 5 mm), skin, breast, and ribs was evaluated to determine the impact of seed placement on the treatment. Results: The overall mean (±SD) difference between implanted and planned positions was 9 ± 5 mm for the aggregate seed population. No significant systematic directional displacements were observed for this whole population. However, for individual patients, systematic displacements were observed, implying that intrapatient offsets occur during the procedure. Mean displacements for seeds in the different anatomic areas were not found to be significantly different from the mean for the entire seed population. However, small directional trends were observed within the anatomy, potentially indicating some bias in the delivery. Despite observed differences between the planned and implanted seed positions, the median (range) V{sub 90} for the 20 patients was 97% (66%-100%), and acceptable dosimetry was achieved for critical structures. Conclusions: No significant trends or systematic errors were observed in the placement of seeds in PBSI, including seeds implanted directly into the seroma. Recorded seed displacements may be related to intrapatient setup adjustments. Despite observed seed

  6. Novel high dose rate lip brachytherapy technique to improve dose homogeneity and reduce toxicity by customized mold

    International Nuclear Information System (INIS)

    Feldman, Jon; Appelbaum, Limor; Sela, Mordechay; Voskoboinik, Ninel; Kadouri, Sarit; Weinberger, Jeffrey; Orion, Itzhak; Meirovitz, Amichay

    2014-01-01

    The purpose of this study is to describe a novel brachytherapy technique for lip Squamous Cell Carcinoma, utilizing a customized mold with embedded brachytherapy sleeves, which separates the lip from the mandible, and improves dose homogeneity. Seven patients with T2 lip cancer treated with a “sandwich” technique of High Dose Rate (HDR) brachytherapy to the lip, consisting of interstitial catheters and a customized mold with embedded catheters, were reviewed for dosimetry and outcome using 3D planning. Dosimetric comparison was made between the “sandwich” technique to “classic” – interstitial catheters only plan. We compared dose volume histograms for Clinical Tumor Volume (CTV), normal tissue “hot spots” and mandible dose. We are reporting according to the ICRU 58 and calculated the Conformal Index (COIN) to show the advantage of our technique. The seven patients (ages 36–81 years, male) had median follow-up of 47 months. Four patients received Brachytherapy and External Beam Radiation Therapy, 3 patients received brachytherapy alone. All achieved local control, with excellent esthetic and functional results. All patients are disease free. The Customized Mold Sandwich technique (CMS) reduced the high dose region receiving 150% (V150) by an average of 20% (range 1–47%), The low dose region (les then 90% of the prescribed dose) improved by 73% in average by using the CMS technique. The COIN value for the CMS was in average 0.92 as opposed to 0.88 for the interstitial catheter only. All differences (excluding the low dose region) were statistically significant. The CMS technique significantly reduces the high dose volume and increases treatment homogeneity. This may reduce the potential toxicity to the lip and adjacent mandible, and results in excellent tumor control, cosmetic and functionality

  7. Time scales of transient enhanced diffusion: Free and clustered interstitials

    Science.gov (United States)

    Cowern, N. E. B.; Huizing, H. G. A.; Stolk, P. A.; Visser, C. C. G.; de Kruif, R. C. M.; Kyllesbech Larsen, K.; Privitera, V.; Nanver, L. K.; Crans, W.

    1996-12-01

    Transient enhanced diffusion (TED) and electrical activation after nonamorphizing Si implantations into lightly B-doped Si multilayers shows two distinct timescales, each related to a different class of interstitial defect. At 700°C, ultrafast TED occurs within the first 15 s with a B diffusivity enhancement of > 2 × 10 5. Immobile clustered B is present at low concentration levels after the ultrafast transient and persists for an extended period (˜ 10 2-10 3 s). The later phase of TED exhibits a near-constant diffusivity enhancement of ≈ 1 × 10 4, consistent with interstitial injection controlled by dissolving {113} interstitial clusters. The relative contributions of the ultrafast and regular TED regimes to the final diffusive broadening of the B profile depends on the proportion of interstitials that escape capture by {113} clusters growing within the implant damage region upon annealing. Our results explain the ultrafast TED recently observed after medium-dose B implantation. In that case there are enough B atoms to trap a large proportion of interstitials in SiB clusters, and the remaining interstitials contribute to TED without passing through an intermediate {113} defect stage. The data on the ultrafast TED pulse allows us to extract lower limits for the diffusivities of the Si interstitial ( DI > 2 × 10 -10 cm 2s -1) and the B interstitial(cy) defect ( DBi > 2 × 10 -13 cm 2s -1) at 700°C.

  8. Salvage high-dose-rate brachytherapy for local prostate cancer recurrence after radical radiotherapy

    Directory of Open Access Journals (Sweden)

    V. A. Solodkiy

    2016-01-01

    Full Text Available Studies salvage interstitial radiation therapy for recurrent prostate cancer, launched at the end of the XX century. In recent years, more and more attention is paid to high-dose-rate brachytherapy (HDR-BT as a method of treating local recurrence.The purpose of research – preliminary clinical results of salvage high-dose-rate brachytherapy applied in cases of suspected local recurrence or of residual tumour after radiotherapy.Preliminary findings indicate the possibility of using HDR-BT, achieving local tumor control with low genitourinary toxicity.

  9. Conventional Versus Automated Implantation of Loose Seeds in Prostate Brachytherapy: Analysis of Dosimetric and Clinical Results

    International Nuclear Information System (INIS)

    Genebes, Caroline; Filleron, Thomas; Graff, Pierre; Jonca, Frédéric; Huyghe, Eric; Thoulouzan, Matthieu; Soulie, Michel; Malavaud, Bernard; Aziza, Richard; Brun, Thomas; Delannes, Martine; Bachaud, Jean-Marc

    2013-01-01

    Purpose: To review the clinical outcome of I-125 permanent prostate brachytherapy (PPB) for low-risk and intermediate-risk prostate cancer and to compare 2 techniques of loose-seed implantation. Methods and Materials: 574 consecutive patients underwent I-125 PPB for low-risk and intermediate-risk prostate cancer between 2000 and 2008. Two successive techniques were used: conventional implantation from 2000 to 2004 and automated implantation (Nucletron, FIRST system) from 2004 to 2008. Dosimetric and biochemical recurrence-free (bNED) survival results were reported and compared for the 2 techniques. Univariate and multivariate analysis researched independent predictors for bNED survival. Results: 419 (73%) and 155 (27%) patients with low-risk and intermediate-risk disease, respectively, were treated (median follow-up time, 69.3 months). The 60-month bNED survival rates were 95.2% and 85.7%, respectively, for patients with low-risk and intermediate-risk disease (P=.04). In univariate analysis, patients treated with automated implantation had worse bNED survival rates than did those treated with conventional implantation (P<.0001). By day 30, patients treated with automated implantation showed lower values of dose delivered to 90% of prostate volume (D90) and volume of prostate receiving 100% of prescribed dose (V100). In multivariate analysis, implantation technique, Gleason score, and V100 on day 30 were independent predictors of recurrence-free status. Grade 3 urethritis and urinary incontinence were observed in 2.6% and 1.6% of the cohort, respectively, with no significant differences between the 2 techniques. No grade 3 proctitis was observed. Conclusion: Satisfactory 60-month bNED survival rates (93.1%) and acceptable toxicity (grade 3 urethritis <3%) were achieved by loose-seed implantation. Automated implantation was associated with worse dosimetric and bNED survival outcomes

  10. Conventional Versus Automated Implantation of Loose Seeds in Prostate Brachytherapy: Analysis of Dosimetric and Clinical Results

    Energy Technology Data Exchange (ETDEWEB)

    Genebes, Caroline, E-mail: genebes.caroline@claudiusregaud.fr [Radiation Oncology Department, Institut Claudius Regaud, Toulouse (France); Filleron, Thomas; Graff, Pierre [Radiation Oncology Department, Institut Claudius Regaud, Toulouse (France); Jonca, Frédéric [Department of Urology, Clinique Ambroise Paré, Toulouse (France); Huyghe, Eric; Thoulouzan, Matthieu; Soulie, Michel; Malavaud, Bernard [Department of Urology and Andrology, CHU Rangueil, Toulouse (France); Aziza, Richard; Brun, Thomas; Delannes, Martine; Bachaud, Jean-Marc [Radiation Oncology Department, Institut Claudius Regaud, Toulouse (France)

    2013-11-15

    Purpose: To review the clinical outcome of I-125 permanent prostate brachytherapy (PPB) for low-risk and intermediate-risk prostate cancer and to compare 2 techniques of loose-seed implantation. Methods and Materials: 574 consecutive patients underwent I-125 PPB for low-risk and intermediate-risk prostate cancer between 2000 and 2008. Two successive techniques were used: conventional implantation from 2000 to 2004 and automated implantation (Nucletron, FIRST system) from 2004 to 2008. Dosimetric and biochemical recurrence-free (bNED) survival results were reported and compared for the 2 techniques. Univariate and multivariate analysis researched independent predictors for bNED survival. Results: 419 (73%) and 155 (27%) patients with low-risk and intermediate-risk disease, respectively, were treated (median follow-up time, 69.3 months). The 60-month bNED survival rates were 95.2% and 85.7%, respectively, for patients with low-risk and intermediate-risk disease (P=.04). In univariate analysis, patients treated with automated implantation had worse bNED survival rates than did those treated with conventional implantation (P<.0001). By day 30, patients treated with automated implantation showed lower values of dose delivered to 90% of prostate volume (D90) and volume of prostate receiving 100% of prescribed dose (V100). In multivariate analysis, implantation technique, Gleason score, and V100 on day 30 were independent predictors of recurrence-free status. Grade 3 urethritis and urinary incontinence were observed in 2.6% and 1.6% of the cohort, respectively, with no significant differences between the 2 techniques. No grade 3 proctitis was observed. Conclusion: Satisfactory 60-month bNED survival rates (93.1%) and acceptable toxicity (grade 3 urethritis <3%) were achieved by loose-seed implantation. Automated implantation was associated with worse dosimetric and bNED survival outcomes.

  11. High-Risk Prostate Cancer With Gleason Score 8–10 and PSA Level ≤15 ng/ mL Treated With Permanent Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Fang, L. Christine; Merrick, Gregory S.; Butler, Wayne M.; Galbreath, Robert W.; Murray, Brian C.; Reed, Joshua L.; Adamovich, Edward; Wallner, Kent E.

    2011-01-01

    Purpose: With widespread prostate-specific antigen (PSA) screening, there has been an increase in men diagnosed with high-risk prostate cancer defined by a Gleason score (GS) ≥8 coupled with a relatively low PSA level. The optimal management of these patients has not been defined. Cause-specific survival (CSS), biochemical progression-free survival (bPFS), and overall survival (OS) were evaluated in brachytherapy patients with a GS ≥8 and a PSA level ≤15 ng/mL with or without androgen-deprivation therapy (ADT). Methods and Materials: From April 1995 to October 2005, 174 patients with GS ≥8 and a PSA level ≤15 ng/mL underwent permanent interstitial brachytherapy. Of the patients, 159 (91%) received supplemental external beam radiation, and 113 (64.9%) received ADT. The median follow-up was 6.6 years. The median postimplant Day 0 minimum percentage of the dose covering 90% of the target volume was 121.1% of prescription dose. Biochemical control was defined as a PSA level ≤0.40 ng/mL after nadir. Multiple parameters were evaluated for impact on survival. Results: Ten-year outcomes for patients without and with ADT were 95.2% and 92.5%, respectively, for CSS (p = 0.562); 86.5% and 92.6%, respectively, for bPFS (p = 0.204); and 75.2% and 66.0%, respectively, for OS (p = 0.179). The median post-treatment PSA level for biochemically controlled patients was <0.02 ng/mL. Multivariate analysis failed to identify any predictors for CSS, whereas bPFS and OS were most closely related to patient age. Conclusions: Patients with GS ≥8 and PSA level ≤15 ng/mL have excellent bPFS and CSS after brachytherapy with supplemental external beam radiotherapy. The use of ADT did not significantly impact bPFS, CSS, or OS.

  12. Investigation on curative efficacy for malignant tumor by implantation '125I permanent brachytherapy seeds

    International Nuclear Information System (INIS)

    Hu Shu; Gao Zhou; Jia Shaowei; Cheng Xianyi; Chen Junhui; Yin Weihua; Sun Desheng

    2011-01-01

    Twenty inpatients suffered from malignant tumors with twenty-four lesions were treated with 125 I permanent brachytherapy seed in Peking University Shenzhen Hospital, and the feasibility, curative effect and adverse effect of the treatment were observed. Before 125 I seeds implantation, the three-dimensional treatment planning was preconcerted. There were two methods to implant 125 I seeds. One was to insert the seeds in the location of residual focus and metastatic lesions of the tumors directly in ordinary operations or through laparoscopy under general anesthesia. The other w as to implant the seeds into the tumors through percutaneous needles by the guidance of CT scanning or color doppler ultrasonography under local anesthesia. The implantations for all of the 20 patients (24 lesions) were performed successfully. During and one week after the implantation, the distributions of the planted seeds were approximately the same as the scheduled three-dimensional treatment planning, and no seed migration was found. Adverse reactions during and after the operation were slight and recovered after correlative treatments. Clinical symptoms were palliated and ser um tumor marker decreased to a different extent among most patients. The complete remission (CR) rate is 20.00% (4/20 patients ), the partial emission (PR) rate is 35.00% (7/20 patients), the stable disease (SD) rate is 30.00% (6/20 patients), the progressive disease (PD) rate is 15.00% (3/20 patients), and the overall response rate (CR + PR) is 53.33% (8 patients). 125 I seeds implantation for targeted therapy is convenient, safe and effective on malignant tumor, and is well worth advanced application. (authors)

  13. On-line implant reconstruction in HDR brachytherapy

    International Nuclear Information System (INIS)

    Kolkman-Deurloo, Inger-Karine K.; Kruijf, Wilhelmus J.M. de; Levendag, Peter C.

    2006-01-01

    Background and purpose: To evaluate the accuracy of on-line planning in an Integrated Brachytherapy Unit (IBU) using dedicated image distortion correction algorithms, correcting the geometric distortion and magnetic distortion separately, and to determine the effect of the reconstruction accuracy on clinical treatment plans in terms of deviations in treatment time and dose. Patients and methods: The reconstruction accuracy has been measured using 20 markers, positioned at well known locations in a QA phantom. Treatment plans of two phantoms representing clinical implant geometries, have been compared with reference plans to determine the effect of the reconstruction accuracy on the treatment plan. Before clinical introduction, treatment plans of three representative patients, based on on-line reconstruction, have been compared with reference plans. Results: The average reconstruction error for 10 in. images reduces from -0.6 mm (range -2.6 to +1.0 mm) to -0.2 mm (range -1.2 to +0.6 mm) after image distortion correction and for 15 in. images from 0.8 mm (range -0.5 to +3.0 mm) to 0.0 mm (range -0.8 to +0.8 mm). The error in case of eccentric positioning of the phantom, i.e. 0.8 mm (range -1.0 to +3.3 mm), reduces to 0.1 mm (range -0.5 to +0.9 mm). Correction of the image distortions reduces the deviation in the calculated treatment time of maximally 2.7% to less than 0.8% in case of eccentrically positioned clinical phantoms. The deviation in the treatment time or reference dose in the plans based on on-line reconstruction with image distortion correction of the three patient examples is smaller than 0.3%. Conclusions: Accurate on-line implant reconstruction using the IBU localiser and dedicated correction algorithms separating the geometric distortion and the magnetic distortion is possible. The results fulfill the minimum requirements as imposed by the Netherlands Commission on Radiation Dosimetry (NCS) without limitations regarding the usable range of the field

  14. Effect of edema, relative biological effectiveness, and dose heterogeneity on prostate brachytherapy

    International Nuclear Information System (INIS)

    Wang, Jian Z.; Mayr, Nina A.; Nag, Subir; Montebello, Joseph; Gupta, Nilendu; Samsami, Nina; Kanellitsas, Christos

    2006-01-01

    Many factors influence response in low-dose-rate (LDR) brachytherapy of prostate cancer. Among them, edema, relative biological effectiveness (RBE), and dose heterogeneity have not been fully modeled previously. In this work, the generalized linear-quadratic (LQ) model, extended to account for the effects of edema, RBE, and dose heterogeneity, was used to assess these factors and their combination effect. Published clinical data have shown that prostate edema after seed implant has a magnitude (ratio of post- to preimplant volume) of 1.3-2.0 and resolves exponentially with a half-life of 4-25 days over the duration of the implant dose delivery. Based on these parameters and a representative dose-volume histogram (DVH), we investigated the influence of edema on the implant dose distribution. The LQ parameters (α=0.15 Gy -1 and α/β=3.1 Gy) determined in earlier studies were used to calculate the equivalent uniform dose in 2 Gy fractions (EUD 2 ) with respect to three effects: edema, RBE, and dose heterogeneity for 125 I and 103 Pd implants. The EUD 2 analysis shows a negative effect of edema and dose heterogeneity on tumor cell killing because the prostate edema degrades the dose coverage to tumor target. For the representative DVH, the V 100 (volume covered by 100% of prescription dose) decreases from 93% to 91% and 86%, and the D 90 (dose covering 90% of target volume) decrease from 107% to 102% and 94% of prescription dose for 125 I and 103 Pd implants, respectively. Conversely, the RBE effect of LDR brachytherapy [versus external-beam radiotherapy (EBRT) and high-dose-rate (HDR) brachytherapy] enhances dose effect on tumor cell kill. In order to balance the negative effects of edema and dose heterogeneity, the RBE of prostate brachytherapy was determined to be approximately 1.2-1.4 for 125 I and 1.3-1.6 for 103 Pd implants. These RBE values are consistent with the RBE data published in the literature. These results may explain why in earlier modeling studies

  15. Ultrasonography-guided cobalt-60 brachytherapy for malignant glioma

    International Nuclear Information System (INIS)

    Sakai, Noboru; Takenaka, Katsunobu; Ueda, Tatsuya

    1989-01-01

    Brachytherapy with cobalt-60 source is reported. In this method it is characterized that the source is inserted interstitially with remote control system by after-loading method via outer catheter (using tandem tube), which was established in the center of residual tumor, using ultrasonography guide with trepanation, or intraoperatively put within the dead space after tumor resection. Six cases of deep-seated and recurrent malignant glioma, were treated with this method. A total dose of 20 to 45 Gy (10 to 15 Gy/day for 2 to 3 days) was delivered to the target. Additionally conventional external irradiation was followed. The effect of cobalt-60 brachytherapy on such tumors were favorable especially for well-circumscribed glioma less than 3 cm on CT scan. (author)

  16. Stereotactic interstitial brachytherapy for the treatment of oligodendroglial brain tumors

    Energy Technology Data Exchange (ETDEWEB)

    El Majdoub, Faycal; Neudorfer, Clemens; Maarouf, Mohammad [University Hospital of Cologne, Department of Stereotaxy and Functional Neurosurgery, Cologne (Germany); University of Witten/Herdecke, Department of Stereotaxy and Functional Neurosurgery, Center of Neurosurgery, Cologne-Merheim Medical Center (CMMC), Cologne (Germany); Blau, Tobias; Deckert, Martina [University Hospital of Cologne, Department of Neuropathology, Cologne (Germany); Hellmich, Martin [University Hospital of Cologne, Institute of Statistics, Informatics and Epidemiology, Cologne (Germany); Buehrle, Christian [University Hospital of Cologne, Department of Stereotaxy and Functional Neurosurgery, Cologne (Germany); Sturm, Volker [University Hospital of Cologne, Department of Stereotaxy and Functional Neurosurgery, Cologne (Germany); University Hospital of Wurzburg, Department of Neurosurgery, Wuerzburg (Germany)

    2015-12-15

    We evaluated the treatment of oligodendroglial brain tumors with interstitial brachytherapy (IBT) using {sup 125}iodine seeds ({sup 125}I) and analyzed prognostic factors. Between January 1991 and December 2010, 63 patients (median age 43.3 years, range 20.8-63.4 years) suffering from oligodendroglial brain tumors were treated with {sup 125}I IBT either as primary, adjuvantly after incomplete resection, or as salvage therapy after tumor recurrence. Possible prognostic factors influencing disease progression and survival were retrospectively investigated. The actuarial 2-, 5-, and 10-year overall and progression-free survival rates after IBT for WHO II tumors were 96.9, 96.9, 89.8 % and 96.9, 93.8, 47.3 %; for WHO III tumors 90.3, 77, 54.9 % and 80.6, 58.4, 45.9 %, respectively. Magnetic resonance imaging demonstrated complete remission in 2 patients, partial remission in 13 patients, stable disease in 17 patients and tumor progression in 31 patients. Median time to progression for WHO II tumors was 87.6 months and for WHO III tumors 27.8 months. Neurological status improved in 10 patients and remained stable in 20 patients, while 9 patients deteriorated. There was no treatment-related mortality. Treatment-related morbidity was transient in 11 patients. WHO II, KPS ≥ 90 %, frontal location, and tumor surface dose > 50 Gy were associated with increased overall survival (p ≤ 0.05). Oligodendroglioma and frontal location were associated with a prolonged progression-free survival (p ≤ 0.05). Our study indicates that IBT achieves local control rates comparable to surgery and radio-/chemotherapy treatment, is minimally invasive, and safe. Due to the low rate of side effects, IBT may represent an attractive option as part of a multimodal treatment schedule, being supplementary to microsurgery or as a salvage therapy after chemotherapy and conventional irradiation. (orig.) [German] Die Behandlung oligodendroglialer Hirntumoren durch die interstitielle Brachytherapie

  17. Conservative treatment of anal canal carcinoma with external radiotherapy and interstitial brachytherapy, with or without chemotherapy: long-term results; Traitement conservateur des cancers du canal anal par irradiation suivie de curietherapie interstitielle, avec ou sans chimiotherapie: resultats a long terme

    Energy Technology Data Exchange (ETDEWEB)

    Berger, C.; Felix-Faure, C.; Chauvet, B.; Vincent, P.; Alfonsi, M.; Coudurier, P.; Plat, F.; Reboul, F. [Clinique Sainte-Catherine, 84 - Avignon (France)

    1999-12-01

    Purpose: a retrospective analysis of conservative treatment of anal canal cancers with external radiation therapy and interstitial brachytherapy with or without chemotherapy. Patients and methods: from 1986 to 1996, 69 patients were treated with external radiotherapy (40 Gy/20 fractions) and interstitial brachytherapy (20 Gy) after a mean interval of six weeks for a localized epidermoid carcinoma of the anal canal. Patients who did not complete the whole therapeutic sequence were not included. Forty-five patients received additional 5-fluorouracil- and/or mitomycin C-based chemotherapy regimen. Results: acute toxicity was acceptable. Complete response rate was 81%. Actuarial local control rate was at two and five years, 65% and 59% respectively (median follow-up: eight years). At two, five and ten years, actuarial colostomy rate was 26%, 33% and 33% respectively, and colostomy-free survival rates 61%, 47% and 37%. Overall survival at two, five and ten years was 81%, 65% and 53% respectively. Distant metastases occurred in 11 patients (16%). Prognostic factors for overall survival were performance status (PS) (79% survival at five years for patients with PS 0 versus 50% for patients with PS 1-3, P = 0.04) and tumor stage (80% at five years for T1-T2 versus 53% for T3-T4, P = 0.03). Overall treatment time less than 12 weeks and time interval between external radiotherapy and brachytherapy inferior than six weeks were associated with a better local control (P = 0.05). In multivariate analysis, these prognostic factors were not significant. Conclusion: these results confirm the efficacy of external radiotherapy and brachytherapy in the treatment of small anal canal cancers, and point out the need for improving treatment outcome of larger tumors. (author)

  18. Seed loss through the urinary tract after prostate brachytherapy: examining the role of cystoscopy and urine straining post implant

    International Nuclear Information System (INIS)

    Stutz, Michael; Petrikas, James; Raslowsky, Michael; Lee, Plato; Gurel, Michelle; Moran, Brian

    2003-01-01

    This study describes one institution's experience with seed retrieval through the urinary tract and makes recommendations for cystoscopy and urine straining post prostate brachytherapy (PB). 1794 patients from two separate cohorts covering different time periods (early versus late) were analyzed. All patients were preplanned with a modified peripheral loading technique and implanted with preloaded needles ( 125 I or 103 Pd) under ultrasound guidance. A catheter was used to delineate the urethra during the volume study but was not used during the implant. All patients underwent post implant cystoscopy. All patients were instructed to strain their urine for seven days post implant and return any seeds to our center. In our experience, seed loss through the urinary tract is a common event after PB, occurring in 29.7% of patients and was more common in patients from the early cohort, those implanted with 125 I seeds or those patients with prior transurethral resection of the prostate. Average seed loss per case, however, represents only 0.58% of total activity. We continue to recommend routine post implant cystoscopy for seed retrieval and periprocedural management. We no longer recommend that patients strain their urine at home after documenting a low rate of seed loss after discharge

  19. High-Dose-Rate Brachytherapy of a Single Implant With Two Fractions Combined With External Beam Radiotherapy for Hormone-Naive Prostate Cancer

    International Nuclear Information System (INIS)

    Sato, Morio; Mori, Takashi; Shirai, Shintaro; Kishi, Kazushi; Inagaki, Takeshi; Hara, Isao

    2008-01-01

    Purpose: To evaluate the preliminary outcomes of high-dose-rate (HDR) brachytherapy of a single implant with two fractions and external beam radiotherapy (EBRT) for hormone-naive prostate cancer. Methods and Materials: Between March 2000 and Sept 2003, a total of 53 patients with tumor Stage T1c-T3b N0 M0 prostate cancer were treated with HDR brachytherapy boost doses (7.5 Gy/fraction) and 50-Gy EBRT during a 5.5-week period. Median follow-up was 61 months. Patients were divided into groups with localized (T1c-T2b) and advanced disease (T3a-T3b). We used the American Society for Therapeutic Radiology and Oncology (ASTRO) definition for biochemical failure. According to recommendations of the Radiation Therapy Oncology Group-ASTRO Phoenix Consensus Conference, biochemical failure-free control rates (BF-FCRs) at 3 years were investigated as 2 years short of the median follow-up. Results: Between April 2000 and Sept 2007, Common Terminology Criteria for Adverse Events Version 2.0 late Grade 2 genitourinary and gastrointestinal toxicity rates were 0% and 3.8%, respectively. Erectile preservation was 25% at 5 years. Overall survival was 88.1% and cause-specific survival was 100%. At 3 years, ASTRO BF-FCRs of the localized and advanced groups were 100% and 42%, respectively (p = 0.001). Conclusions: The HDR brachytherapy of a single implant with two fractions plus EBRT is effective in treating patients with localized hormone-naive prostate cancer, with the least genitourinary and gastrointestinal toxicities; however, longer median BF-FCR follow-up is required to assess these findings

  20. Ejaculatory Function After Permanent 125I Prostate Brachytherapy for Localized Prostate Cancer

    International Nuclear Information System (INIS)

    Huyghe, Eric; Delannes, Martine; Wagner, Fabien M.; Delaunay, Boris; Nohra, Joe; Thoulouzan, Matthieu; Shut-Yee, J. Yeung; Plante, Pierre; Soulie, Michel; Thonneau, Patrick; Bachaud, Jean Marc

    2009-01-01

    Purpose: Ejaculatory function is an underreported aspect of male sexuality in men treated for prostate cancer. We conducted the first detailed analysis of ejaculatory function in patients treated with permanent 125 I prostate brachytherapy for localized prostate cancer. Patients and Methods: Of 270 sexually active men with localized prostate cancer treated with permanent 125 I prostate brachytherapy, 241 (89%), with a mean age of 65 years (range, 43-80), responded to a mailed questionnaire derived from the Male Sexual Health Questionnaire regarding ejaculatory function. Five aspects of ejaculatory function were examined: frequency, volume, dry ejaculation, pleasure, and pain. Results: Of the 241 sexually active men, 81.3% had conserved ejaculatory function after prostate brachytherapy; however, the number of patients with rare/absent ejaculatory function was double the pretreatment number (p < .0001). The latter finding was correlated with age (p < .001) and the preimplant International Index of Erectile Function score (p < .001). However, 84.9% of patients with maintained ejaculatory function after implantation reported a reduced volume of ejaculate compared with 26.9% before (p < .001), with dry ejaculation accounting for 18.7% of these cases. After treatment, 30.3% of the patients experienced painful ejaculation compared with 12.9% before (p = .0001), and this was associated with a greater number of implanted needles (p = .021) and the existence of painful ejaculation before implantation (p < .0001). After implantation, 10% of patients who continued to be sexually active experienced no orgasm compared with only 1% before treatment. in addition, more patients experienced late/difficult or weak orgasms (p = .001). Conclusion: Most men treated with brachytherapy have conserved ejaculatory function after prostate brachytherapy. However, most of these men experience a reduction in volume and a deterioration in orgasm.

  1. An analysis of intraoperative versus post-operative dosimetry with CT, CT-MRI fusion and XMR for the evaluation of permanent prostate brachytherapy implants

    International Nuclear Information System (INIS)

    Acher, Peter; Puttagunta, Srikanth; Rhode, Kawal; Morris, Stephen; Kinsella, Janette; Gaya, Andrew; Dasgupta, Prokar; Deehan, Charles; Beaney, Ronald; Popert, Rick; Keevil, Stephen

    2010-01-01

    Background and purpose: To assess the agreement between intraoperative and post-operative dosimetry and to identify factors that influence dose calculations of prostate brachytherapy implants. Materials and methods: Patients treated with prostate brachytherapy implants underwent post-operative CT and XMR (combined X-ray and MR) imaging. Dose-volume histograms were calculated from CT, XMR and CT-MR fusion data and compared with intraoperative values for two observers. Multiple linear regression models assessed the influences of intraoperative D90, gland oedema, gland volume, source loss and migration, and implanted activity/volume prostate on post-operative D90. Results: Forty-nine patients were studied. The mean D90 differences (95% confidence limits) between intraoperative and post-operative CT, XMR and CT-MR fusion assessments were: 11 Gy (-22, 45), 18 Gy (-13, 49) and 20 Gy (-17, 58) for Observer 1; and 15 Gy (-34, 63), 13 Gy (-29, 55) and 14 Gy (-27, 54) for Observer 2. Multiple linear regression modelling showed that the observed oedema and intraoperative D90 were significant independent variables for the prediction of post-operative D90 values for both observers using all modalities. Conclusion: This is the first study to report Bland-Altman agreement analysis between intraoperative and post-operative dosimetry. Agreement is poor. Post-operative dosimetry is dependent on the intraoperative D90 and the subjectively outlined gland volume.

  2. Novel prostate brachytherapy technique: Improved dosimetric and clinical outcome

    International Nuclear Information System (INIS)

    Nobes, Jenny P.; Khaksar, Sara J.; Hawkins, Maria A.; Cunningham, Melanie J.; Langley, Stephen E.M.; Laing, Robert W.

    2008-01-01

    Purpose: Erectile dysfunction following prostate brachytherapy is reported to be related to dose received by the penile bulb. To minimise this, whilst preserving prostate dosimetry, we have developed a technique for I-125 seed brachytherapy using both stranded seeds and loose seeds delivered with a Mick applicator, and implanted via the sagittal plane on trans-rectal ultrasound. Materials and methods: Post-implant dosimetry and potency rates were compared in 120 potent patients. In Group 1, 60 patients were treated using a conventional technique of seeds implanted in a modified-uniform distribution. From January 2005, a novel technique was developed using stranded seeds peripherally and centrally distributed loose seeds implanted via a Mick applicator (Group 2). The latter technique allows greater flexibility when implanting the seeds at the apex. Each patient was prescribed a minimum peripheral dose of 145 Gy. No patients received external beam radiotherapy or hormone treatment. There was no significant difference in age or pre-implant potency score (mean IIEF-5 score 22.4 vs. 22.6, p = 0.074) between the two groups. Results: The new technique delivers lower penile bulb doses (D 25 as %mPD - Group 1: 61.2 ± 35.7, Group 2: 29.7 ± 16.0, p 50 as %mPD - Group 1: 45.8 ± 26.9, Group 2: 21.4 ± 11.7, p 90 - Group 1: 147 Gy ± 21.1, Group 2: 155 Gy ± 16.7, p = 0.03). At 2 years, the potency rate was also improved: Group 1: 61.7%; Group 2: 83.3% (p = 0.008). Conclusions: In this study, the novel brachytherapy technique using both peripheral stranded seeds and central loose seeds delivered via a Mick applicator results in a lower penile bulb dose whilst improving prostate dosimetry, and may achieve higher potency rates

  3. The American brachytherapy society survey of brachytherapy practice for carcinoma of the cervix in the United States.

    Science.gov (United States)

    Nag, S; Orton, C; Young, D; Erickson, B

    1999-04-01

    The purpose of this study was to survey the brachytherapy practice for cervical cancer in the United States. The Clinical Research Committee of the American Brachytherapy Society (ABS) performed a retrospective survey of individual physicians of the ABS and American Society of Therapeutic Radiologists and Oncologists regarding the details of the brachytherapy techniques they personally used in the treatment of cervical cancer patients for the year 1995. The replies (some of which may have been an estimate only) were tabulated. The scope of this survey did not allow us to verify the data by chart audits. A total of about 3500 questionnaires were mailed out; 521 responses were received. Of these responders, 206 (40%) did not perform any brachytherapy for carcinoma of the cervix in 1995. Of the other 315 responders reporting a total of 4892 patients treated in 1995, 88% used low dose rate (LDR) while 24% used high dose rate (HDR). There was a wide variation in the doses used. For LDR treatments, the median total external beam radiation therapy (EBRT) dose was 45 and 50 Gy and the LDR dose was 42 and 45 Gy for early and advanced cancers, respectively. For HDR treatments, the median EBRT dose was 48 and 50 Gy and the median HDR dose was 29 and 30 Gy for early and advanced cancers, respectively. The median dose per fraction was 6 Gy for a median of five fractions. Interstitial brachytherapy was used as a component of the treatment in 6% of the patients by 21% of responders. Very few responders treated with pulsed or medium dose rates. This retrospective survey showed the current brachytherapy practice pattern in the treatment of cervical cancer in the United States and can serve as a basis for future prospective national brachytherapy data registry. There was wide variation in the practice pattern, emphasizing the urgent need for consensus on these issues. Copyright 1999 Academic Press.

  4. Dose calculation in brachytherapy with microcomputers

    International Nuclear Information System (INIS)

    Elbern, A.W.

    1989-01-01

    The computer algorithms, that allow the calculation of brachytherapy doses and its graphic representation for implants, using programs developed for Pc microcomputers are presented. These algorithms allow to localized the sources in space, from their projection in radiographics images and trace isodose counter. (C.G.C.) [pt

  5. Impact of prostate edema on cell survival and tumor control after permanent interstitial brachytherapy for early stage prostate cancers

    Science.gov (United States)

    Chen, Zhe (Jay); Roberts, Kenneth; Decker, Roy; Pathare, Pradip; Rockwell, Sara; Nath, Ravinder

    2011-01-01

    Previous studies have shown that the procedure-induced prostate edema during permanent interstitial brachytherapy (PIB) can cause significant variations in the dose delivered to the prostate gland. Because the clinical impact of edema-induced dose variations depends strongly on the magnitude of the edema, the temporal pattern of its resolution and its interplay with the decay of radioactivity and the underlying biological processes of tumor cells (such as tumor potential doubling time), we investigated the impact of edema-induced dose variations on the tumor cell survival and tumor control probability after PIB with the 131Cs, 125I and 103Pd sources used in current clinical practice. The exponential edema resolution model reported by Waterman et al. (Int. J. Radiat. Oncol. Biol. Phys. 41, 1069–1077–1998) was used to characterize the edema evolutions observed previously during clinical PIB for prostate cancer. The concept of biologically effective dose (BED), taking into account tumor cell proliferation and sublethal damage repair during dose delivery, was used to characterize the effects of prostate edema on cell survival and tumor control probability. Our calculation indicated that prostate edema, if not taken into account appropriately, can increase the cell survival and decrease the probability of local control of PIB. The edema-induced increase in cell survival increased with increasing edema severity, decreasing half-life for radioactive decay and decreasing energy of the photons energy emitted by the source. At the doses currently prescribed for PIB and for prostate cancer cells characterized by nominal radiobiology parameters recommended by AAPM TG-137, PIB using 125I sources was less affected by edema than PIB using 131Cs or 103Pd sources due to the long radioactive decay half-life of 125I. The effect of edema on PIB using 131Cs or 103Pd was similar. The effect of edema on 103Pd PIB was slightly greater, even though the decay half-life of 103Pd (17 days

  6. Critical Organ Preservation in Reirradiation Brachytherapy by Injectable Spacer

    International Nuclear Information System (INIS)

    Kishi, Kazushi; Sonomura, Tetsuo; Shirai, Shintaro; Sato, Morio; Tanaka, Kayo

    2009-01-01

    Purpose: This case series study evaluated the feasibility and effectiveness of an interstitial high-dose rate brachytherapy (HDR-BT) procedure combined with an at-risk organ-sparing procedure. Methods and Materials: Thirty patients who were scheduled for reirradiation treatment for recurrent cancer after receiving a median dose of 60 Gy (range, 44-70 Gy) in 2-Gy fractions of previous external beam treatment were enrolled. Thirteen patients had lesions in the head and neck, and other lesions were located in the axilla, skeleton, breast, pelvis, and abdominal wall. Chief complaints included local masses (for 25) and refractory pain (for 21). After high-dose rate brachytherapy applicator needle implantation, an optimal CT-based three-dimensional brachytherapy plan was created with a virtual at-risk organ shift from the target. According to the plan, hyaluronic acid gel was injected to maintain the shift during irradiation. The prescribed dose was the result of an individualized tradeoff between target dose and at-risk organ dose, to avoid serious complications. A single-fraction dose of 18.0 Gy (median, equivalent to 75.6 Gy at an α/β value of 3; range, 16-20 Gy) was applied to the tumor. Results: The at-risk organ dose decreased from 9.1 ± 0.9 Gy to 4.4 ± 0.4 Gy (mean ± standard deviation, p < 0.01), and the normal tissue complication probability decreased from 60.8% ± 12.6% to 16.1% ± 19.8% (p < 0.01). The shift effect lasted at least 4 hours and disappeared gradually. Distinct tumor shrinkage in 20 of 21 eligible patients, including tumor disappearance in 6 patients, pain reduction in 18 of 21 eligible patients, and no unexpected late toxicity greater than grade 2 were observed during the 19.5-month observation period. Conclusions: This at-risk organ-sparing preservation procedure may provide a safe and efficient reirradiation treatment.

  7. External beam techniques to boost cervical cancer when brachytherapy is not an option—theories and applications

    Science.gov (United States)

    Kilic, Sarah; Khan, Atif J.; Beriwal, Sushil; Small, William

    2017-01-01

    The management of locally advanced cervical cancer relies on brachytherapy (BT) as an integral part of the radiotherapy delivery armamentarium. Occasionally, intracavitary BT is neither possible nor available. In these circumstances, post-external beam radiotherapy (EBRT) interstitial brachytherapy and/or hysterectomy may represent viable options that must be adequately executed in a timely manner. However, if these options are not applicable due to patient related or facility related reasons, a formal contingency plan should be in place. Innovative EBRT techniques such as intensity modulated and stereotactic radiotherapy may be considered for patients unable to undergo brachytherapy. Relying on provocative arguments and recent data, this review explores the rationale for and limitations of non-brachytherapy substitutes in that setting aiming to establish a formal process for the optimal execution of this alternative plan. PMID:28603722

  8. Timing of computed tomography-based postimplant assessment following permanent transperineal prostate brachytherapy

    International Nuclear Information System (INIS)

    Prestidge, Bradley R.; Bice, William S.; Kiefer, Eric J.; Prete, James J.

    1998-01-01

    Purpose: To establish the rate of resolution of prostatic edema following transperineal interstitial permanent prostate brachytherapy, and to determine the results and impact of timing of the postimplant assessment on the dose-volume relationship. Methods and Materials: A series of 19 consecutive patients with early-stage adenocarcinoma of the prostate receiving transperineal interstitial permanent prostate brachytherapy, were enrolled in this study. Twelve received 125 I and seven received 103 Pd. Postoperative assessment included a computed tomographic (CT) scan on postoperative days 1, 8, 30, 90, and 180. On each occasion, CT scans were performed on a GE helical unit at 3-mm abutting slices, 15-cm field of view. Prostate volumes were outlined on CT scans by a single clinician. Following digitization of the volumes and radioactive sources, volumes and dose-volume histograms were calculated. The prostate volume encompassed by the 80% and 100% reference isodose volumes was calculated. Results: Preimplant transrectal ultrasound determined volumes varied from 17.5 to 38.6 cc (median 27.9 cc). Prostate volumes previously defined on 40 randomly selected postimplant CT scans were compared in a blinded fashion to a second CT-derived volume and ranged from -32% to +24%. The Pearson correlation coefficient for prostate CT volume reproducibility was 0.77 (p < 0.03). CT scan-determined volume performed on postoperative day 1 was an average of 41.4% greater than the volume determined by preimplant ultrasound. Significant decreases in average volume were seen during the first month postoperatively. Average volume decreased 14% from day 1 to day 8, 10% from day 8 to day 30, 3% from day 30 to day 90, and 2% thereafter. Coverage of the prostate volume by the 80% isodose volume increased from 85.6% on postoperative day 1 to 92.2% on postoperative day 180. The corresponding increase in the 100% reference dose coverage of the prostate volume ranged from 73.1% to 83.3% between

  9. Development of an adjoint sensitivity field-based treatment-planning technique for the use of newly designed directional LDR sources in brachytherapy.

    Science.gov (United States)

    Chaswal, V; Thomadsen, B R; Henderson, D L

    2012-02-21

    The development and application of an automated 3D greedy heuristic (GH) optimization algorithm utilizing the adjoint sensitivity fields for treatment planning to assess the advantage of directional interstitial prostate brachytherapy is presented. Directional and isotropic dose kernels generated using Monte Carlo simulations based on Best Industries model 2301 I-125 source are utilized for treatment planning. The newly developed GH algorithm is employed for optimization of the treatment plans for seven interstitial prostate brachytherapy cases using mixed sources (directional brachytherapy) and using only isotropic sources (conventional brachytherapy). All treatment plans resulted in V100 > 98% and D90 > 45 Gy for the target prostate region. For the urethra region, the D10(Ur), D90(Ur) and V150(Ur) and for the rectum region the V100cc, D2cc, D90(Re) and V90(Re) all are reduced significantly when mixed sources brachytherapy is used employing directional sources. The simulations demonstrated that the use of directional sources in the low dose-rate (LDR) brachytherapy of the prostate clearly benefits in sparing the urethra and the rectum sensitive structures from overdose. The time taken for a conventional treatment plan is less than three seconds, while the time taken for a mixed source treatment plan is less than nine seconds, as tested on an Intel Core2 Duo 2.2 GHz processor with 1GB RAM. The new 3D GH algorithm is successful in generating a feasible LDR brachytherapy treatment planning solution with an extra degree of freedom, i.e. directionality in very little time.

  10. Development of an adjoint sensitivity field-based treatment-planning technique for the use of newly designed directional LDR sources in brachytherapy

    Science.gov (United States)

    Chaswal, V.; Thomadsen, B. R.; Henderson, D. L.

    2012-02-01

    The development and application of an automated 3D greedy heuristic (GH) optimization algorithm utilizing the adjoint sensitivity fields for treatment planning to assess the advantage of directional interstitial prostate brachytherapy is presented. Directional and isotropic dose kernels generated using Monte Carlo simulations based on Best Industries model 2301 I-125 source are utilized for treatment planning. The newly developed GH algorithm is employed for optimization of the treatment plans for seven interstitial prostate brachytherapy cases using mixed sources (directional brachytherapy) and using only isotropic sources (conventional brachytherapy). All treatment plans resulted in V100 > 98% and D90 > 45 Gy for the target prostate region. For the urethra region, the D10Ur, D90Ur and V150Ur and for the rectum region the V100cc, D2cc, D90Re and V90Re all are reduced significantly when mixed sources brachytherapy is used employing directional sources. The simulations demonstrated that the use of directional sources in the low dose-rate (LDR) brachytherapy of the prostate clearly benefits in sparing the urethra and the rectum sensitive structures from overdose. The time taken for a conventional treatment plan is less than three seconds, while the time taken for a mixed source treatment plan is less than nine seconds, as tested on an Intel Core2 Duo 2.2 GHz processor with 1GB RAM. The new 3D GH algorithm is successful in generating a feasible LDR brachytherapy treatment planning solution with an extra degree of freedom, i.e. directionality in very little time.

  11. Deformable registration of x-ray to MRI for post-implant dosimetry in prostate brachytherapy

    Science.gov (United States)

    Park, Seyoun; Song, Danny Y.; Lee, Junghoon

    2016-03-01

    Post-implant dosimetric assessment in prostate brachytherapy is typically performed using CT as the standard imaging modality. However, poor soft tissue contrast in CT causes significant variability in target contouring, resulting in incorrect dose calculations for organs of interest. CT-MR fusion-based approach has been advocated taking advantage of the complementary capabilities of CT (seed identification) and MRI (soft tissue visibility), and has proved to provide more accurate dosimetry calculations. However, seed segmentation in CT requires manual review, and the accuracy is limited by the reconstructed voxel resolution. In addition, CT deposits considerable amount of radiation to the patient. In this paper, we propose an X-ray and MRI based post-implant dosimetry approach. Implanted seeds are localized using three X-ray images by solving a combinatorial optimization problem, and the identified seeds are registered to MR images by an intensity-based points-to-volume registration. We pre-process the MR images using geometric and Gaussian filtering. To accommodate potential soft tissue deformation, our registration is performed in two steps, an initial affine transformation and local deformable registration. An evolutionary optimizer in conjunction with a points-to-volume similarity metric is used for the affine registration. Local prostate deformation and seed migration are then adjusted by the deformable registration step with external and internal force constraints. We tested our algorithm on six patient data sets, achieving registration error of (1.2+/-0.8) mm in < 30 sec. Our proposed approach has the potential to be a fast and cost-effective solution for post-implant dosimetry with equivalent accuracy as the CT-MR fusion-based approach.

  12. The surgical viability and radiological monitoring of brain implants of bioactive micro-seeds in an animal model

    International Nuclear Information System (INIS)

    Silva, Giane X.O.; Campos, Tarcisio Passos Ribeiro de; Siqueira, Savio Lana; Maciel, Marcelo B.

    2005-01-01

    The interstitial implant is a therapeutic modality in brachytherapy of the head and neck. Presently, the seeds implanted in tumors in the central nervous system are metallic I-125. After the full emission of the radionuclide, the seed remains inert in the implanted area. Bioactive ceramic seeds have been prepared for this research group incorporating Sm-152 to be active in Sm-153. The main goal of the present study is the development of a the surgical technique for implanting the biodegradable radioactive micro-seeds in the brains of rabbits, as well as the observation of the clinical reactions of the animal after implantation of two sets of three seeds. The surgical procedure consisted of performing two separate perforations 10 mm from each other in the skull, permitting the implantation of two groups of three seeds, totaling six seeds. The results of the pilot study showed the effectiveness of the surgical procedure and of the biocompatibility of the seeds and the lack of presence of adverse reactions, functional sequels, or inflammation in a follow up 50 days post-surgery. Such seeds of reduced volume, 0.2 x 1.6 mm, could be monitored by computerized tomography 30 days after implanting. (author)

  13. Perioperative Interstitial High-Dose-Rate Brachytherapy for the Treatment of Recurrent Keloids

    DEFF Research Database (Denmark)

    Jiang, Ping; Baumann, René; Dunst, Jürgen

    2016-01-01

    PURPOSE: To prospectively evaluate high-dose-rate brachytherapy in the treatment of therapy-resistant keloids and report first results, with emphasis on feasibility and early treatment outcome. METHODS AND MATERIALS: From 2009 to 2014, 24 patients with 32 recurrent keloids were treated with immed...

  14. Pulsed dose rate brachytherapy – is it the right way?

    Directory of Open Access Journals (Sweden)

    Janusz Skowronek

    2010-10-01

    Full Text Available Pulsed dose rate (PDR-BT treatment is a brachytherapy modality that combines physical advantages of high-doserate (HDR-BT technology (isodose optimization, radiation safety with the radiobiological advantages of low-dose-rate (LDR-BT brachytherapy. Pulsed brachytherapy consists of using stronger radiation source than for LDR-BT and producing series of short exposures of 10 to 30 minutes in every hour to approximately the same total dose in the sameoverall time as with the LDR-BT. Modern afterloading equipment offers certain advantages over interstitial or intracavitaryinsertion of separate needles, tubes, seeds or wires. Isodose volumes in tissues can be created flexibly by a combinationof careful placement of the catheter and the adjustment of the dwell times of the computerized stepping source.Automatic removal of the radiation sources into a shielded safe eliminates radiation exposures to staff and visitors.Radiation exposure is also eliminated to the staff who formerly loaded and unloaded multiplicity of radioactive sources into the catheters, ovoids, tubes etc. This review based on summarized clinical investigations, analyses the feasibility and the background to introduce this brachytherapy technique and chosen clinical applications of PDR-BT.

  15. 1.5 T conventional MR-guided iodine-125 interstitial implants for hepatocellular carcinoma: Feasibility and preliminary clinical experience

    International Nuclear Information System (INIS)

    Lin Zhengyu; Lin Jun; Lin Cong; Li Yinguan; Chen Shaoming; Hu Jianping; Hu Rui; Chen Jin

    2012-01-01

    Purpose: To evaluate the feasibility and therapeutic efficacy of 1.5 T conventional MR-guided percutaneous interstitial implantation of I-125 radioactive seeds in patients with hepatocellular carcinoma (HCC). Materials and methods: The institutional ethics committee approved this study. After imformed consent was obtained, twenty-three patients suffering from a total of 65 HCC lesions were treated with I-125 seed permanent interstitial implantation under the guidance of a 1.5 T conventional magnetic resonance imaging (MRI) system. The FSE T2WI, T1 FSPGR, FIESTA 2D, 3D Dyn T1WI sequences were used to guide an 18G MR-compatible needle inserted into the lesions to introduce the I-125 seeds. The response evaluation criteria in solid tumors (RECIST) were used to evaluate the curative effect. Results: The needle and I-125 seed were seen clearly on MRI images. The final dose delivered to total decay was 173.46 ± 32.44 Gy (range, 110–270 Gy) as calculated by postoperative TPS. The complete response (CR) was seen in 22 lesions (33.8%), partial response (PR) in 24 lesions (36.9%), stable disease (SD) in 9 lesions (13.8%), and progressive disease (PD) in 10 lesions (15.4%). The post-operative alpha-fetoprotein (AFP) was decreased (t = 3.117, P = 0.005 < 0.05). Two patients were observed a small area of subcapsular bleeding that did not lead to any symptoms or clinical sequelae. Conclusion: MR-guided I-125 implantation for HCC is technically feasible and effective.

  16. A survey of current clinical practice in permanent and temporary prostate brachytherapy: 2010 update.

    Science.gov (United States)

    Buyyounouski, Mark K; Davis, Brian J; Prestidge, Bradley R; Shanahan, Thomas G; Stock, Richard G; Grimm, Peter D; Demanes, D Jeffrey; Zaider, Marco; Horwitz, Eric M

    2012-01-01

    To help establish patterns of care and standards of care of interstitial permanent low-dose-rate (LDR) and temporary high-dose-rate brachytherapy for prostate cancer and to compare the results with a similar 1998 American Brachytherapy Society (ABS) survey. A comprehensive questionnaire intended to survey specific details of current clinical brachytherapy practice was provided to the participants of the seventh ABS Prostate Brachytherapy School. Responses were tabulated and descriptive statistics are reported. Sixty-five brachytherapy practitioners responded to the survey. Eighty-nine percent (89%) of respondents performed LDR and 49% perform high-dose-rate brachytherapy. The median number of years of experience for LDR brachytherapists increased from 5 to 10 years over the course of the 12 years since the preceding survey. Compared with the first ABS, a smaller proportion of respondents received formal brachytherapy residency training (43% vs. 56%) or formal "hands-on" brachytherapy training (15% vs. 63%). There has been a marked decline in the utilization of the Mick applicator (Mick Radio-Nuclear Instruments, Inc., Mount Vernon, NY, USA) (60% vs. 28%) and an increase in the use of stranded seeds (40% vs. 11%). Compliance with postimplant dosimetry was higher in the 2010 survey. This survey does suggest an evolution in the practice of LDR brachytherapy since 1998 and aids in identifying aspects that require further progress or investigation. ABS guidelines and other practice recommendations appear to impact the practice of brachytherapy. Copyright © 2012 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  17. Quality of Life of Oral Cancer Patients After Low-Dose-Rate Interstitial Brachytherapy

    International Nuclear Information System (INIS)

    Yoshimura, Ryo-ichi; Shibuya, Hitoshi; Miura, Masahiko; Watanabe, Hiroshi; Ayukawa, Fumio; Hayashi, Keiji; Toda, Kazuma

    2009-01-01

    Purpose: To assess the quality of life (QOL) of oral cancer patients treated with low-dose-rate interstitial brachytherapy (LDR-BT) alone. Methods and Materials: Between June 2005 and July 2006, a total of 56 patients with oral cancer were enrolled in this prospective study. QOL was assessed by means of the core questionnaire and head and neck questionnaire module of the European Organization for Research and Treatment of Cancer (EORTC Quality of Life Questionnaire-Core 30 [QLQ-C30] and QLQ Head and Neck 35 [H and N35]). The questionnaires were distributed to the patients before the start of treatment and 3 months, 6 months, and 12 months after the start of LDR-BT. Results: It was possible to analyze the results for 20 of the initial 56 patients because they did not experience metastasis or recurrence during this study. No functions or symptoms asked about in the QLQ-C30 deteriorated during the first year. The emotional function score steadily and significantly increased. No symptoms in the QLQ-H and N35 significantly deteriorated. The scores for pain, trouble with social eating, and weight loss on the QLQ-H and N35 steadily and significantly decreased. Age, gender, and LDR-BT source had no effect on the change in QOL during the first year, but T-stage significantly affected the change in global health status, tumor site affected the changes in swallowing, sensory problems, sticky saliva, and complications affected the changes in pain, swallowing, and mouth opening. Conclusions: QOL of oral cancer patients treated with LDR-BT is high. However, tumor stage, tumor site, and complications affected the changes in a few functions and symptoms during the first year

  18. Correlation of conventional simulation x-ray films and CT images for HDR-brachytherapy catheters reconstruction

    International Nuclear Information System (INIS)

    Rajendran, M.; Reddy, K.D.; Reddy, R.M.; Reddy, J.M.; Reddy, B.V.N.; Kiran Kumar; Gopi, S.; Dharaniraj; Janardhanan

    2002-01-01

    In order to plan a brachytherapy implant, it is imperative that implant reconstruction is done accurately. The purpose of this paper is to evaluate whether implant reconstruction done with transverse CT images is comparable to reconstruction done with conventional x-ray films

  19. Three dimensional implementation of anisotropy corrected fast fourier transform dose calculation around brachytherapy seeds

    International Nuclear Information System (INIS)

    Kyeremeh, P.O.

    2011-01-01

    Current-available brachytherapy dose computation algorithms ignore heterogeneities such as tissue-air interfaces, shielded gynaecological colpostats, and tissue-composition variations in source implants despite dose computation errors as large as 40%. A convolution kernel, which takes into consideration anisotropy of the dose distribution around a brachytherapy source, and to compute dose in the presence of tissue and applicator heterogeneities, has been established. Resulting from the convolution kernel are functions with polynomial and exponential terms. the solution to the convolution integral was represented by the Fast Fourier transform. The Fast Fourier transform has shown enough potency in accounting for errors due to these heterogeneities and the versatility of this Fast Fourier transform is evident from its capability of switching in between fields. Thus successful procedures in external beam could be adopted in brachytherapy to a yield similar effect. A dose deposition kernel was developed for a 64x64x64 matrix size with wrap around ordering and convoluted with the distribution of the sources in 3D. With MatLab's inverse Fast Fourier transform, dose rate distribution for a given array of interstitial sources, typical of brachytherapy was calculated. The shape of the dose rate distribution peaks appeared comparable with the output expected from computerized treatment planning systems for brachytherapy. Subsequently, the study confirmed the speed and accuracy of dose computation using the FFT convolution as well juxtaposed. Although, dose rate peaks from both the FFT convolution and the TPS(TG43) did not compare quantitatively, which was mainly due to the TPS(TG43) initiation computations from the origin (0,0,0) unlike the FFT convolution which uses sampling points; N=1,2,3..., there is a strong basis for establishing parity since the dose rate peaks compared qualitatively. With both modes compared, the discrepancies in the dose rates ranged between 3.6% to

  20. Perioperative Interstitial High-Dose-Rate Brachytherapy for the Treatment of Recurrent Keloids: Feasibility and Early Results

    Energy Technology Data Exchange (ETDEWEB)

    Jiang, Ping, E-mail: ping.jiang@uksh.de [Department of Radiation Oncology, University Clinic Schleswig-Holstein, Campus Kiel, Kiel (Germany); Baumann, René [Department of Radiation Oncology, University Clinic Schleswig-Holstein, Campus Kiel, Kiel (Germany); Dunst, Juergen [Department of Radiation Oncology, University Clinic Schleswig-Holstein, Campus Kiel, Kiel (Germany); Department of Radiation Oncology, University of Copenhagen, Copenhagen (Denmark); Geenen, Matthias [Department of Reconstructive Surgery, Lubinus Clinic Kiel, Kiel (Germany); Siebert, Frank-André [Department of Radiation Oncology, University Clinic Schleswig-Holstein, Campus Kiel, Kiel (Germany); Niehoff, Peter [Department of Radiation Oncology, University Clinic Schleswig-Holstein, Campus Kiel, Kiel (Germany); Department of Radiation Oncology, Community Clinic Köln, Köln (Germany); Department of Radiation Oncology, University Witten/Herdecke, Witten (Germany); Bertolini, Julia; Druecke, Daniel [Department of Reconstructive Surgery, University Clinic Schleswig-Holstein, Campus Kiel, Kiel (Germany)

    2016-03-01

    Purpose: To prospectively evaluate high-dose-rate brachytherapy in the treatment of therapy-resistant keloids and report first results, with emphasis on feasibility and early treatment outcome. Methods and Materials: From 2009 to 2014, 24 patients with 32 recurrent keloids were treated with immediate perioperative high-dose-rate brachytherapy; 3 patients had been previously treated with adjuvant external beam radiation therapy and presented with recurrences in the pretreated areas. Two or more different treatment modalities had been tried in all patients and had failed to achieve remission. After (re-)excision of the keloids, a single brachytherapy tube was placed subcutaneously before closing the wound. The target volume covered the scar in total length. Brachytherapy was given in 3 fractions with a single dose of 6 Gy in 5 mm tissue depth. The first fraction was given within 6 hours after surgery, the other 2 fractions on the first postoperative day. Thus, a total dose of 18 Gy in 3 fractions was administered within 36 hours after the resection. Results: The treatment was feasible in all patients. No procedure-related complications (eg, secondary infections) occurred. Nineteen patients had keloid-related symptoms before treatment like pain and pruritus; disappearance of symptoms was noticed in all patients after treatment. After a median follow-up of 29.4 months (range, 7.9-72.4 months), 2 keloid recurrences and 2 mildly hypertrophied scars were observed. The local control rate was 94%. Pigmentary abnormalities were detected in 3 patients, and an additional 6 patients had a mild delay in the wound-healing process. Conclusions: The early results of this study prove the feasibility and the efficacy of brachytherapy for the prevention of keloids. The results also suggest that brachytherapy may be advantageous in the management of high-risk keloids or as salvage treatment for failure after external beam therapy.

  1. The impact of prostate edema on cell survival and tumor control after permanent interstitial brachytherapy for early stage prostate cancers

    International Nuclear Information System (INIS)

    Chen Zhe; Roberts, Kenneth; Decker, Roy; Pathare, Pradip; Rockwell, Sara; Nath, Ravinder

    2011-01-01

    Previous studies have shown that procedure-induced prostate edema during permanent interstitial brachytherapy (PIB) can cause significant variations in the dose delivered to the prostate gland. Because the clinical impact of edema-induced dose variations strongly depends on the magnitude of the edema, the temporal pattern of its resolution and its interplay with the decay of radioactivity and the underlying biological processes of tumor cells (such as tumor potential doubling time), we investigated the impact of edema-induced dose variations on the tumor cell survival and tumor control probability after PIB with the 131 Cs, 125 I and 103 Pd sources used in current clinical practice. The exponential edema resolution model reported by Waterman et al (1998 Int. J. Radiat. Oncol. Biol. Phys. 41 1069-77) was used to characterize the edema evolutions previously observed during clinical PIB for prostate cancer. The concept of biologically effective dose, taking into account tumor cell proliferation and sublethal damage repair during dose delivery, was used to characterize the effects of prostate edema on cell survival and tumor control probability. Our calculation indicated that prostate edema, if not appropriately taken into account, can increase the cell survival and decrease the probability of local control of PIB. The magnitude of an edema-induced increase in cell survival increased with increasing edema severity, decreasing half-life of radioactive decay and decreasing photon energy emitted by the source. At the doses currently prescribed for PIB and for prostate cancer cells characterized by nominal radiobiology parameters recommended by AAPM TG-137, PIB using 125 I sources was less affected by edema than PIB using 131 Cs or 103 Pd sources due to the long radioactive decay half-life of 125 I. The effect of edema on PIB using 131 Cs or 103 Pd was similar. The effect of edema on 103 Pd PIB was slightly greater, even though the decay half-life of 103 Pd (17 days) is

  2. Use of brachytherapy with permanent implants of iodine-125 in localized prostate cancer; La curietherapie par implants permanents d'I-125 dans le cancer localise de la prostate

    Energy Technology Data Exchange (ETDEWEB)

    Bladou, F.; Serment, G. [Hopital Salvador, Service d' Urologie, 13 - Marseille (France); Salem, N.; Simonian, M. [Hopital Salvador, Dept. de Radiotherapie, 13 - Marseille (France); Rosello, R.; Ternier, F. [Institut Paoli-Calmettes, Dept. de Radiologie, 13 - Marseille (France)

    2002-07-01

    Approximately 15,000 cases of early stage prostate cancer T1 and T2 are diagnosed every year in France by testing for PSA and performing prostatic biopsies. The treatment of these localized forms is based in most cases on radical prostatectomy or nn external beam radiotherapy. Although the ontological results obtained by these two therapeutic methods are satisfactory and equivalent in the long term, the side effects can be important. For a number of years, trans-perineal brachytherapy using permanent implants of iodine -125 or palladium-103 has proved itself as an alternative therapy with equivalent medium to long-term results. The low urinary, digestive and sexual side effects of prostate brachytherapy are important reasons for the enthusiasm among patients and the medical community for this therapy and the growing number of applications and centres which practice it. In September 1998 we started the prostate brachytherapy programmes- in Marseilles with close collaboration between the department of urology of the Hopital Salvator, and the departments of radiotherapy, medical imaging and medical physics of the Institut Paoli-Calmettes. To date, around 250 patients with localized adenocarcinoma of the prostate have benefited from this alternative therapy in our centre. Preliminary results, with a 3 year-follow-up, are comparable to results published in the literature by pioneer teams. (authors)

  3. Independent technique of verifying high-dose rate (HDR) brachytherapy treatment plans

    International Nuclear Information System (INIS)

    Saw, Cheng B.; Korb, Leroy J.; Darnell, Brenda; Krishna, K. V.; Ulewicz, Dennis

    1998-01-01

    Purpose: An independent technique for verifying high-dose rate (HDR) brachytherapy treatment plans has been formulated and validated clinically. Methods and Materials: In HDR brachytherapy, dwell times at respective dwell positions are computed, using an optimization algorithm in a HDR treatment-planning system to deliver a specified dose to many target points simultaneously. Because of the variability of dwell times, concerns have been expressed regarding the ability of the algorithm to compute the correct dose. To address this concern, a commercially available low-dose rate (LDR) algorithm was used to compute the doses at defined distances, based on the dwell times obtained from the HDR treatment plans. The percent deviation between doses computed using the HDR and LDR algorithms were reviewed for HDR procedures performed over the last year. Results: In this retrospective study, the difference between computed doses using the HDR and LDR algorithms was found to be within 5% for about 80% of the HDR procedures. All of the reviewed procedures have dose differences of less than 10%. Conclusion: An independent technique for verifying HDR brachytherapy treatment plans has been validated based on clinical data. Provided both systems are available, this technique is universal in its applications and not limited to either a particular implant applicator, implant site, or implant type

  4. Dosimetric and radiobiological comparison of volumetric modulated arc therapy, high-dose rate brachytherapy, and low-dose rate permanent seeds implant for localized prostate cancer

    Energy Technology Data Exchange (ETDEWEB)

    Yang, Ruijie, E-mail: ruijyang@yahoo.com; Zhao, Nan; Liao, Anyan; Wang, Hao; Qu, Ang

    2016-10-01

    To investigate the dosimetric and radiobiological differences among volumetric modulated arc therapy (VMAT), high-dose rate (HDR) brachytherapy, and low-dose rate (LDR) permanent seeds implant for localized prostate cancer. A total of 10 patients with localized prostate cancer were selected for this study. VMAT, HDR brachytherapy, and LDR permanent seeds implant plans were created for each patient. For VMAT, planning target volume (PTV) was defined as the clinical target volume plus a margin of 5 mm. Rectum, bladder, urethra, and femoral heads were considered as organs at risk. A 78 Gy in 39 fractions were prescribed for PTV. For HDR and LDR plans, the dose prescription was D{sub 90} of 34 Gy in 8.5 Gy per fraction, and 145 Gy to clinical target volume, respectively. The dose and dose volume parameters were evaluated for target, organs at risk, and normal tissue. Physical dose was converted to dose based on 2-Gy fractions (equivalent dose in 2 Gy per fraction, EQD{sub 2}) for comparison of 3 techniques. HDR and LDR significantly reduced the dose to rectum and bladder compared with VMAT. The D{sub mean} (EQD{sub 2}) of rectum decreased 22.36 Gy in HDR and 17.01 Gy in LDR from 30.24 Gy in VMAT, respectively. The D{sub mean} (EQD{sub 2}) of bladder decreased 6.91 Gy in HDR and 2.53 Gy in LDR from 13.46 Gy in VMAT. For the femoral heads and normal tissue, the mean doses were also significantly reduced in both HDR and LDR compared with VMAT. For the urethra, the mean dose (EQD{sub 2}) was 80.26, 70.23, and 104.91 Gy in VMAT, HDR, and LDR brachytherapy, respectively. For localized prostate cancer, both HDR and LDR brachytherapy were clearly superior in the sparing of rectum, bladder, femoral heads, and normal tissue compared with VMAT. HDR provided the advantage in sparing of urethra compared with VMAT and LDR.

  5. Dosimetric and radiobiological comparison of volumetric modulated arc therapy, high-dose rate brachytherapy, and low-dose rate permanent seeds implant for localized prostate cancer

    International Nuclear Information System (INIS)

    Yang, Ruijie; Zhao, Nan; Liao, Anyan; Wang, Hao; Qu, Ang

    2016-01-01

    To investigate the dosimetric and radiobiological differences among volumetric modulated arc therapy (VMAT), high-dose rate (HDR) brachytherapy, and low-dose rate (LDR) permanent seeds implant for localized prostate cancer. A total of 10 patients with localized prostate cancer were selected for this study. VMAT, HDR brachytherapy, and LDR permanent seeds implant plans were created for each patient. For VMAT, planning target volume (PTV) was defined as the clinical target volume plus a margin of 5 mm. Rectum, bladder, urethra, and femoral heads were considered as organs at risk. A 78 Gy in 39 fractions were prescribed for PTV. For HDR and LDR plans, the dose prescription was D 90 of 34 Gy in 8.5 Gy per fraction, and 145 Gy to clinical target volume, respectively. The dose and dose volume parameters were evaluated for target, organs at risk, and normal tissue. Physical dose was converted to dose based on 2-Gy fractions (equivalent dose in 2 Gy per fraction, EQD 2 ) for comparison of 3 techniques. HDR and LDR significantly reduced the dose to rectum and bladder compared with VMAT. The D mean (EQD 2 ) of rectum decreased 22.36 Gy in HDR and 17.01 Gy in LDR from 30.24 Gy in VMAT, respectively. The D mean (EQD 2 ) of bladder decreased 6.91 Gy in HDR and 2.53 Gy in LDR from 13.46 Gy in VMAT. For the femoral heads and normal tissue, the mean doses were also significantly reduced in both HDR and LDR compared with VMAT. For the urethra, the mean dose (EQD 2 ) was 80.26, 70.23, and 104.91 Gy in VMAT, HDR, and LDR brachytherapy, respectively. For localized prostate cancer, both HDR and LDR brachytherapy were clearly superior in the sparing of rectum, bladder, femoral heads, and normal tissue compared with VMAT. HDR provided the advantage in sparing of urethra compared with VMAT and LDR.

  6. Brachytherapy for head and neck cancer. Treatment results and future prospect

    International Nuclear Information System (INIS)

    Shibuya, Hitoshi; Yoshimura, Ro-ichi; Miura, Masahiko; Ayukawa, Fumio; Watanabe, Hiroshi

    2005-01-01

    Following the increasing desire of many patients to keep the form and function of speech and swallowing, interstitial brachytherapy has become the main treatment for head and neck cancer. In addition, aged and physically handicapped patients who are refused general anesthesia have come to be referred to our clinic to receive less invasive and curative treatment. In the field of brachytherapy for head and neck cancers, less complicated and more superior treatment results have been achieved following the introduction of spacers, computer dosimetry and so on. As a result of these efforts, treatment results have come to fulfill the desire of patients and their families. During the past 43 years from 1962 to 2005, we have treated over 2, 100 patients of head and neck cancer including 850 with stage I·II oral tongue carcinoma by brachytherapy and acquired a lot of important and precious data including the treatment results, multiple primary cancers as well as radiation-induced cancers. (author)

  7. Intravascular brachytherapy for peripheral vascular disease

    Directory of Open Access Journals (Sweden)

    Hagen, Anja

    2008-09-01

    Full Text Available Scientific background: Percutaneous transluminal angioplasties (PTA through balloon dilatation with or without stenting, i.e. vessel expansion through balloons with or without of implantation of small tubes, called stents, are used in the treatment of peripheral artery occlusive disease (PAOD. The intravascular vessel irradiation, called intravascular brachytherapy, promises a reduction in the rate of repeated stenosis (rate of restenosis after PTA. Research questions: The evaluation addresses questions on medical efficacy, cost-effectiveness as well as ethic, social and legal implications in the use of brachytherapy in PAOD patients. Methods: A systematic literature search was conducted in August 2007 in the most important medical electronic databases for publications beginning from 2002. The medical evaluation included randomized controlled trials (RCT. The information synthesis was performed using meta-analysis. Health economic modeling was performed with clinical assumptions derived from the meta-analysis and economical assumptions derived from the German Diagnosis Related Groups (G-DRG-2007. Results: Medical evaluation: Twelve publications about seven RCT on brachytherapy vs. no brachytherapy were included in the medical evaluation. Two RCT showed a significant reduction in the rate of restenosis at six and/or twelve months for brachytherapy vs. no brachytherapy after successful balloon dilatation, the relative risk in the meta-analysis was 0.62 (95% CI: 0.46 to 0.84. At five years, time to recurrence of restenosis was significantly delayed after brachytherapy. One RCT showed a significant reduction in the rate of restenosis at six months for brachytherapy vs. no brachytherapy after PTA with optional stenting, the relative risk in the meta-analysis was 0.76 (95% CI: 0.61 to 0.95. One RCT observed a significantly higher rate of late thrombotic occlusions after brachytherapy in the subgroup of stented patients. A single RCT for brachytherapy

  8. Computed tomography-guided interstitial high dose rate brachytherapy for centrally located liver tumours: a single institution study

    Energy Technology Data Exchange (ETDEWEB)

    Tselis, Nikolaos; Chatzikonstantinou, Georgios; Zamboglou, Nikolaos [Klinikum Offenbach, Department of Radiation Oncology, Offenbach am Main (Germany); Kolotas, Christos [Hirslanden Medical Center, Institute for Radiotherapy, Aarau (Switzerland); Milickovic, Natasa; Baltas, Dimos [Klinikum Offenbach, Department of Medical Physics and Engineering, Offenbach am Main (Germany)

    2013-08-15

    To evaluate the clinical outcome of computed tomography (CT)-guided interstitial (IRT) high-dose-rate (HDR) brachytherapy (BRT) in the treatment of unresectable primary and secondary liver malignancies. This report updates and expands our previously described experience with this treatment technique. Forty-one patients with 50 tumours adjacent to the liver hilum and bile duct bifurcation were treated in 59 interventions of CT-guided IRT HDR BRT. The tumours were larger than 4 cm with a median volume of 84 cm{sup 3} (38-1,348 cm{sup 3}). The IRT HDR BRT delivered a median total physical dose of 20.0 Gy (7.0-32.0 Gy) in twice daily fractions of median 7.0 Gy (4.0-10.0 Gy) in 19 patients and in once daily fractions of median 8.0 Gy (7.0-14.0 Gy) in 22 patients. With a median follow-up of 12.4 months, the local control for metastatic hepatic tumours was 89 %, 73 % and 63 % at 6, 12 and 18 months respectively. The local control for primary hepatic tumours was 90 %, 81 % and 50 % at 6, 12 and 18 months respectively. Severe side effects occurred in 5.0 % of interventions with no treatment-related deaths. CT-guided IRT HDR BRT is a promising procedure for the radiation treatment of centrally located liver malignancies. (orig.)

  9. The american brachytherapy society recommendations for permanent prostate brachytherapy postimplant dosimetric analysis

    International Nuclear Information System (INIS)

    Nag, Subir; Bice, William; Wyngaert, Keith de; Prestidge, Bradley; Stock, Richard; Yu Yan

    2000-01-01

    Purpose: The purpose of this report is to establish guidelines for postimplant dosimetric analysis of permanent prostate brachytherapy. Methods: Members of the American Brachytherapy Society (ABS) with expertise in prostate dosimetry evaluation performed a literature review and supplemented with their clinical experience formulated guidelines for performing and analyzing postimplant dosimetry of permanent prostate brachytherapy. Results: The ABS recommends that postimplant dosimetry should be performed on all patients undergoing permanent prostate brachytherapy for optimal patient care. At present, computed tomography (CT)-based dosimetry is recommended, based on availability cost and the ability to image the prostate as well as the seeds. Additional plane radiographs should be obtained to verify the seed count. Until the ideal postoperative interval for CT scanning has been determined, each center should perform dosimetric evaluation of prostate implants at a consistent postoperative interval. This interval should be reported. Isodose displays should be obtained at 50%, 80%, 90%, 100%, 150%, and 200% of the prescription dose and displayed on multiple cross-sectional images of the prostate. A dose-volume histogram (DVH) of the prostate should be performed and the D 90 (dose to 90% of the prostate gland) reported by all centers. Additionally, the D 80, D 100, the fractional V 80, V 90, V 100, V 150, and V 200, (i.e., the percentage of prostate volume receiving 80%, 90%, 100%, 150%, and 200% of the prescribed dose, respectively), the rectal, and urethral doses should be reported and ultimately correlated with clinical outcome in the research environment. On-line real-time dosimetry, the effects of dose heterogeneity, and the effects of tissue heterogeneity need further investigation. Conclusion: It is essential that postimplant dosimetry should be performed on all patients undergoing permanent prostate brachytherapy. Guidelines were established for the performance

  10. Permanent LDR implants in treatment of prostate cancer

    International Nuclear Information System (INIS)

    Skowronek, J.; Kanikowski, M.; Chichel, A.; Zwierzchowski, G.

    2009-01-01

    Low-dose rate brachytherapy (LDR-BT) is a radiation method known for several years in the treatment of localized prostate cancer. The main idea of this method is to implant small radioactive seeds directly into the prostate gland. LDR brachytherapy is applied as a monotherapy and also used along with external beam radiation therapy (EBRT) as a boost. In most cases it is used as a sole radical treatment modality, but not as a palliative treatment. The application of permanent seed implants is a curative treatment alternative in patients with organ- confined cancer, without extracapsular extension of the tumour. This technique is particularly popular in the United States. In Europe, however, high-dose rate brachytherapy (HDR-BT) is more popular in early-stage prostate cancer treatment (as a boost). The aim of this publication is to describe methods, indications, complications and selected results of prostate cancer LDR brachytherapy. (authors)

  11. Rectourethral fistula following LDR brachytherapy.

    Science.gov (United States)

    Borchers, Holger; Pinkawa, Michael; Donner, Andreas; Wolter, Timm P; Pallua, Norbert; Eble, Michael J; Jakse, Gerhard

    2009-01-01

    Modern LDR brachytherapy has drastically reduced rectal toxicity and decreased the occurrence of rectourethral fistulas to <0.5% of patients. Therefore, symptoms of late-onset sequelae are often ignored initially. These fistulas cause severe patient morbidity and require interdisciplinary treatment. We report on the occurrence and management of a rectourethral fistula which occurred 4 years after (125)I seed implantation. Copyright 2009 S. Karger AG, Basel.

  12. Brachytherapy in Europe: philosophies, current practice and future directions

    International Nuclear Information System (INIS)

    Haworth, A.

    2000-01-01

    Full text: Five months sabbatical leave provided an opportunity to visit six radiotherapy centres in France, Holland and England. While brachytherapy philosophies and practices within each country were similar, there were considerable differences in attitudes between countries. The Institute Gustave Roussy, home of the Paris System and host for the French sector confirmed that the Paris System is still very much the preferred dosimetry method in this part of the world. Though their preference for low dose rate brachytherapy is still evident, high dose rate brachytherapy has found some applications but the rules of the Paris System are never far away and the words 'what about the hyperdose sleeve' are firmly implanted into this visitor's brain. The use of real time dosimetry for I-125 prostate brachytherapy at the Institute Curie (Paris) provided an interesting contrast to the standard pre and post implant dosimetry techniques commonly employed elsewhere. The two Dutch centres on the itinerary, in stark contrast to the traditional techniques seen in France, have applied the power of computers to investigate optimisation of the classic dosimetry systems and called on the analysis techniques (DVH, NTCP, TCP etc) now familiar to us all in external beam therapy. The Cookridge Hospital in England fitted somewhere between the French and Dutch centres. This centre showed how both modern and traditional techniques could be applied in an efficient way for a large variety of treatment sites. Copyright (2000) Australasian College of Physical Scientists and Engineers in Medicine

  13. Description of new techniques on interstitial brachytherapy in head and neck (oral cavity)

    International Nuclear Information System (INIS)

    Marsiglia, H.

    1990-01-01

    Several case of epidermoid carcinomas in buccal cavity treated by brachytherapy associated with others therapeutic techniques are described, including the results using different therapeutical methods and the clinical and technical parameters considered on the evolution of the disease. (C.G.C.)

  14. Interstitial irradiation for craniopharyngioma

    International Nuclear Information System (INIS)

    Barlas, O.; Bayindir, C.; Can, M.

    2000-01-01

    The results of interstitial irradiation treatment for craniopharyngioma in two patients with six year follow-ups are presented. Stereotactic interstitial irradiation with iodine-125 sources as sole therapy was employed in two adult patients who refused surgical resection. The diagnoses were confirmed by stereotactic biopsy. The first tumour which underwent interstitial irradiation was solid and 4 cm in diameter, and the second, 2.7 cm in diameter, had both cystic and solid components. The implanted iodine-125 seeds delivered 67 Gy and 60 Gy to tumour periphery at the rate of 12 and 14 cGy/h, respectively, were removed at the end of designated radiation periods. Tumour shrinkage and central hypo density, first observed 3 months after irradiation, continued until one tumour shrank to less than 1 cm at 12 months, and the other disappeared completely at 24 months. In both cases functional integrity was restored, and neither radiation induced toxicity nor recurrence has occurred six years after treatment. The results in these two cases suggest that solid craniopharyngiomas are sensitive to interstitial irradiation. (author)

  15. A method to incorporate interstitial components into the TPS gynecologic rigid applicator library

    Directory of Open Access Journals (Sweden)

    Antonio Otal

    2017-01-01

    Full Text Available Purpose: T2 magnetic resonance imaging (MRI is recommended as the imaging modality for image-guided brachytherapy. In locally advanced cervical carcinoma, combined endocavitary and interstitial applicators are appropriate (Vienna or Utrecht. To cover extensive disease, Template Benidorm (TB was developed. Treatment planning system applicator libraries are currently unavailable for the Utrecht applicator or the TB. The purpose of this work is to develop an applicator library for both applicators. Material and methods: The library developed in this work has been used in the Oncentra Brachytherapy TPS, version 4.3.0, which has a brachytherapy module that includes a library of rigid applicators. To add the needles of the Utrecht applicator and to model the TB, we used FreeCAD and MeshLab. The reconstruction process was based on the points that the rigid section and the interstitial part have in common. This, together with the free length, allowed us to ascertain the position of the tip. Results: In case of the Utrecht applicator, one of the sources of uncertainty in the reconstruction was determining the distance of the tip of needle from the ovoid. In case of the TB, the large number of needles involved made their identification time consuming. The developed library resolved both issues. Conclusions : The developed library for the Utrecht and TB is feasible and efficient improving accuracy. It allows all the required treatment planning to proceed using just a T2 MRI sequence. The additional use of specific free available software applications makes it possible to add this information to the already existing library of the Oncentra Brachytherapy TPS. Specific details not included on this manuscript will be available under request. This library is also currently being implemented also into the Sagiplan v 2.0 TPS.

  16. Brachytherapy for prostate cancer: Comparative characteristics of procedures

    Directory of Open Access Journals (Sweden)

    S. V. Kanaev

    2015-01-01

    Full Text Available The introduction of interstitial radiation sources is the «youngest» of the radical method of treatment of patients with prostate cancer (PC. The high level of efficiency comparable to prostatectomy at a significantly lower rate of complications causes rapid growth of clinical use of brachytherapy (BT. Depending on the radiation source and the mode of administration into the prostate gland are two types BT – high-dose rate (temporary (HDR-BT and low-dose rate (permanent (LDR-BT brachytherapy. At the heart of these two methods are based on a single principle of direct effect of the quantum gamma radiation on the area of interest. However, the differences between the characteristics of isotopes used and technical aspects of the techniques cause the difference in performance and complication rates for expression HDR-BT and LDR-BT.

  17. Assessment of I-125 seed implant accuracy when using the live-planning technique for low dose rate prostate brachytherapy

    Directory of Open Access Journals (Sweden)

    Moorrees Joshua

    2012-11-01

    Full Text Available Abstract Background Low risk prostate cancers are commonly treated with low dose rate (LDR brachytherapy involving I-125 seeds. The implementation of a ‘live-planning’ technique at the Royal Adelaide Hospital (RAH in 2007 enabled the completion of the whole procedure (i.e. scanning, planning and implant in one sitting. ‘Live-planning’ has the advantage of a more reliable delivery of the planned treatment compared to the ‘traditional pre-plan’ technique (where patient is scanned and planned in the weeks prior to implant. During live planning, the actual implanted needle positions are updated real-time on the treatment planning system and the dosimetry is automatically recalculated. The aim of this investigation was to assess the differences and clinical relevance between the planned dosimetry and the updated real-time implant dosimetry. Methods A number of 162 patients were included in this dosimetric study. A paired t-test was performed on the D90, V100, V150 and V200 target parameters and the differences between the planned and implanted dose distributions were analysed. Similarly, dosimetric differences for the organs at risk (OAR were also evaluated. Results Small differences between the primary dosimetric parameters for the target were found. Still, the incidence of hotspots was increased with approximately 20% for V200. Statistically significant increases were observed in the doses delivered to the OAR between the planned and implanted data; however, these increases were consistently below 3% thus probably without clinical consequences. Conclusions The current study assessed the accuracy of prostate implants with I-125 seeds when compared to initial plans. The results confirmed the precision of the implant technique which RAH has in place. Nevertheless, geographical misses, anatomical restrictions and needle displacements during implant can have repercussions for centres without live-planning option if dosimetric changes are not

  18. Simulation of measurement absorbed dose on prostate brachytherapy with radius of prostate 2 cm using MCNP5 with seed implant model isoaid AdvantageTM IAPd-103A

    International Nuclear Information System (INIS)

    Poundra Setiawan; Suharyana; Riyatun

    2015-01-01

    Simulation of measurement absorbed dose on prostate brachytherapy with radius of prostate 2 cm using MCNP5 with seed implant model IsoAid Advantage TM IAPd-103A has been conducted. 103 Pd used as a radioactive source in the seed implant and it has energy gamma emission 20,8 keV with half live 16,9 days and has activity 4 mCi. The prostate cancer is modeled with spherical and it has radius 3 cm, after planting the seed implant 103 Pdover 24,4 days, prostate cancer has absorbed dose 2,172Gy. Lethal dose maximum use 103 Pd is 125 Gy and it was reached with 59 seeds. (author)

  19. Vaginal tolerance of CT based image-guided high-dose rate interstitial brachytherapy for gynecological malignancies

    International Nuclear Information System (INIS)

    Murakami, Naoya; Inaba, Koji; Morota, Madoka; Ito, Yoshinori; Itami, Jun; Kasamatsu, Takahiro; Sumi, Minako; Yoshimura, Ryoichi; Harada, Ken; Kitaguchi, Mayuka; Sekii, Shuhei; Takahashi, Kana; Yoshio, Kotaro

    2014-01-01

    Purpose of this study was to identify predictors of vaginal ulcer after CT based three-dimensional image-guided high-dose-rate interstitial brachytherapy (HDR-ISBT) for gynecologic malignancies. Records were reviewed for 44 female (14 with primary disease and 30 with recurrence) with gynecological malignancies treated with HDR-ISBT with or without external beam radiation therapy. The HDR-ISBT applicator insertion was performed with image guidance by trans-rectal ultrasound and CT. The median clinical target volume was 35.5 ml (2.4-142.1 ml) and the median delivered dose in equivalent dose in 2 Gy fractions (EQD 2 ) for target volume D 90 was 67.7 Gy (48.8-94.2 Gy, doses of external-beam radiation therapy and brachytherapy were combined). For re-irradiation patients, median EQD 2 of D 2cc for rectum and bladder, D 0.5cc , D 1cc , D 2cc , D 4cc , D 6cc and D 8cc for vaginal wall was 91.1 Gy, 100.9 Gy, 260.3 Gy, 212.3 Gy, 170.1 Gy, 117.1 Gy, 105.2 Gy, and 94.7 Gy, respectively. For those without prior radiation therapy, median EQD 2 of D 2cc for rectum and bladder, D 0.5cc , D 1cc , D 2cc , D 4cc , D 6cc and D 8cc for vaginal wall was 56.3 Gy, 54.3 Gy, 147.4 Gy, 126.2 Gy, 108.0 Gy, 103.5 Gy, 94.7 Gy, and 80.7 Gy, respectively. Among five patients with vaginal ulcer, three had prior pelvic radiation therapy in their initial treatment and three consequently suffered from fistula formation. On univariate analysis, re-irradiation and vaginal wall D 2cc in EQD 2 was the clinical predictors of vaginal ulcer (p = 0.035 and p = 0.025, respectively). The ROC analysis revealed that vaginal wall D 2cc is the best predictor of vaginal ulcer. The 2-year incidence rates of vaginal ulcer in the patients with vaginal wall D 2cc in EQD 2 equal to or less than 145 Gy and over 145 Gy were 3.7% and 23.5%, respectively, with a statistically significant difference (p = 0.026). Re-irradiation and vaginal D 2cc is a significant predictor of vaginal ulcer after HDR-ISBT for gynecologic

  20. Validation of a novel robot-assisted 3DUS system for real-time planning and guidance of breast interstitial HDR brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Poulin, Eric; Beaulieu, Luc, E-mail: Luc.Beaulieu@phy.ulaval.ca [Département de Physique, de Génie Physique et d’optique et Centre de Recherche sur le Cancer de l’Université Laval, Université Laval, Québec, Québec G1V 0A6, Canada and Département de Radio-oncologie et Axe Oncologie du Centre de Recherche du CHU de Québec, CHU de Québec, 11 Côte du Palais, Québec, Québec G1R 2J6 (Canada); Gardi, Lori; Barker, Kevin; Montreuil, Jacques; Fenster, Aaron [Imaging Research Laboratories, Robarts Research Institute, 100 Perth Drive, London, Ontario N6A 5K8 (Canada)

    2015-12-15

    Purpose: In current clinical practice, there is no integrated 3D ultrasound (3DUS) guidance system clinically available for breast brachytherapy. In this study, the authors present a novel robot-assisted 3DUS system for real-time planning and guidance of breast interstitial high dose rate (HDR) brachytherapy treatment. Methods: For this work, a new computer controlled robotic 3DUS system was built to perform a hybrid motion scan, which is a combination of a 6 cm linear translation with a 30° rotation at both ends. The new 3DUS scanner was designed to fit on a modified Kuske assembly, keeping the current template grid configuration but modifying the frame to allow the mounting of the 3DUS system at several positions. A finer grid was also tested. A user interface was developed to perform image reconstruction, semiautomatic segmentation of the surgical bed as well as catheter reconstruction and tracking. A 3D string phantom was used to validate the geometric accuracy of the reconstruction. The volumetric accuracy of the system was validated with phantoms using magnetic resonance imaging (MRI) and computed tomography (CT) images. In order to accurately determine whether 3DUS can effectively replace CT for treatment planning, the authors have compared the 3DUS catheter reconstruction to the one obtained from CT images. In addition, in agarose-based phantoms, an end-to-end procedure was performed by executing six independent complete procedures with both 14 and 16 catheters, and for both standard and finer Kuske grids. Finally, in phantoms, five end-to-end procedures were performed with the final CT planning for the validation of 3DUS preplanning. Results: The 3DUS acquisition time is approximately 10 s. A paired Student t-test showed that there was no statistical significant difference between known and measured values of string separations in each direction. Both MRI and CT volume measurements were not statistically different from 3DUS volume (Student t-test: p > 0

  1. A comparison of complications between ultrasound-guided prostate brachytherapy and open prostate brachytherapy

    International Nuclear Information System (INIS)

    Benoit, Ronald M.; Naslund, Michael J.; Cohen, Jeffrey K.

    2000-01-01

    Purpose: Prostate brachytherapy has reemerged during the 1990s as a treatment for clinically localized prostate cancer. The renewed popularity of prostate brachytherapy is largely due to the use of transrectal ultrasound of the prostate, which allows for more accurate isotope placement within the prostate when compared to the open approach. The present study investigates whether this improved cancer control is at the expense of increased morbidity by comparing the morbidity after transrectal ultrasound-guided prostate brachytherapy to the morbidity after prostate brachytherapy performed via an open approach. Methods and Materials: All men in the Medicare population who underwent prostate brachytherapy in the year 1991 were identified. These men were further stratified into those men who underwent prostate brachytherapy via an open approach and the men who underwent prostate brachytherapy with ultrasound guidance. All subsequent inpatient, outpatient, and physician (Part B) Medicare claims for these men from the years 1991-1993 were then analyzed to determine outcomes. Results: In the year 1991, 2124 men in the Medicare population underwent prostate brachytherapy. An open approach was used in 715 men (33.7%), and ultrasound guidance was used in 1409 men (66.3%). Mean age for both cohorts was 73.7 years with a range of 50.7-92.8 years for the ultrasound group and 60.6-92.1 years for the open group. A surgical procedure for the relief of bladder outlet obstruction was performed in 122 men (8.6%) in the ultrasound group and in 54 men (7.6%) in the open group. An artificial urinary sphincter was placed in 2 men (0.14%) in the ultrasound group and in 2 men (0.28%) in the open group. A penile prosthesis was implanted in 10 men (0.71%) in the ultrasound group and in 4 men (0.56%) in the open group. A diagnosis code for urinary incontinence was carried by 95 men (6.7%) in the ultrasound group and by 45 men (6.3%) in the open group. A diagnosis code for erectile dysfunction

  2. SU-F-BRA-03: Integrating Novel Electromagnetic Tracking Hollow Needle Assistance in Permanent Implant Brachytherapy Procedures

    Energy Technology Data Exchange (ETDEWEB)

    Racine, E; Hautvast, G; Binnekamp, D [Philips Group Innovation - Biomedical Systems, Eindhoven (Netherlands); Beaulieu, L [Centre Hospitalier Univ de Quebec, Quebec, QC (Canada)

    2015-06-15

    Purpose: To report on the results of a complete permanent implant brachytherapy procedure assisted by an electromagnetic (EM) hollow needle possessing both 3D tracking and seed drop detection abilities. Methods: End-to-end in-phantom EM-assisted LDR procedures were conducted. The novel system consisted of an EM tracking apparatus (NDI Aurora V2, Planar Field Generator), a 3D US scanner (Philips CX50), a hollow needle prototype allowing 3D tracking and seed drop detection and a specially designed treatment planning software (Philips Healthcare). A tungsten-doped 30 cc spherical agarose prostate immersed in gelatin was used for the treatment. A cylindrical shape of 0.8 cc was carved along its diameter to mimic the urethra. An initial plan of 26 needles and 47 seeds was established with the system. The plan was delivered with the EM-tracked hollow needle, and individual seed drop locations were recorded on the fly. The phantom was subsequently imaged with a CT scanner from which seed positions and contour definitions were obtained. The DVHs were then independently recomputed and compared with those produced by the planning system, both before and after the treatment. Results: Of the 47 seeds, 45 (96%) were detected by the EM technology embedded in the hollow needle design. The executed plan (from CT analysis) differed from the initial plan by 2%, 14% and 8% respectively in terms of V100, D90 and V150 for the prostate, and by 8%, 7% and 10% respectively in terms of D5, V100 and V120 for the urethra. Conclusion: The average DVH deviations between initial and executed plans were within a 5% tolerance imposed for this proof-of-concept assessment. This relatively good concordance demonstrates the feasibility and potential benefits of combining EM tracking and seed drop detection for real-time dosimetry validation and assistance in permanent implant brachytherapy procedures. This project has been entirely funded by Philips Healthcare.

  3. Interstitial Fe in MgO

    CERN Document Server

    Mølholt, T E; Gunnlaugsson, H P; Svane, A; Masenda, H; Naidoo, D; Bharuth-Ram, K; Fanciulli, M; Gislason, H P; Johnston, K; Langouche, G; Ólafsson, S; Sielemann, R; Weyer, G

    2014-01-01

    Isolated Fe-57 atoms were studied in MgO single-crystals by emission Mossbauer spectroscopy following implantation of Mn-57 decaying to Fe-57. Four Mossbauer spectral components were found corresponding to different Fe lattice positions and/or charge states. Two components represent Fe atoms substituting Mg as Fe2+ and Fe3+, respectively; a third component is due to Fe in a strongly implantation-induced disturbed region. The fourth component, which is the focus of this paper, can be assigned to Fe at an interstitial site. Comparison of its measured isomer shift with ab initio calculations suggests that the interstitial Fe is located on, or close to, the face of the rock-salt MgO structure. To harmonize such an assignment with the measured near-zero quadrupole interaction a local motion process (cage motion) of the Fe has to be stipulated. The relation of such a local motion as a starting point for long range diffusion is discussed.

  4. Fat necrosis in women with early-stage breast cancer treated with accelerated partial breast irradiation (APBI) using interstitial brachytherapy

    International Nuclear Information System (INIS)

    Budrukkar, Ashwini; Jagtap, Vikas; Kembhavi, Seema; Munshi, Anusheel; Jalali, Rakesh; Seth, Tanuja; Parmar, Vani; Raj Upreti, Ritu; Badwe, Rajendra; Sarin, Rajiv

    2012-01-01

    Purpose: To report the incidence of clinical, pathological and radiological fat necrosis (FN) in women treated with accelerated partial breast irradiation (APBI) using interstitial brachytherapy (BRT) for early-stage breast cancer and to study certain variables associated with it. Methods and materials: Between May 2000 and August 2008, 171 women were treated with APBI using high dose rate (HDR) BRT. Patients were treated to a dose of 34 Gy/10 fractions/1 week with two fractions/day after intraoperative/postoperative placement of catheters. Results: At a median follow up of 48 months (SD: 28) 20 women developed FN with median time to detection being 24 months (range: 4–62 months, SD: 20). Actuarial 5 and 7 year FN rate was 18% and 23%, respectively. Grade 1 FN was seen in 4, grade 2 in 8 and grade 4 in 8 women. Additional investigations such as aspiration/biopsy were done in 9 patients. Volume of excision was the only significant factor affecting FN (p = 0.04). Conclusions: Actuarial FN rate of 18% at 5 years in our study was comparable to other reported series of FN. Median time of detection of FN was 24 months. Higher volume of excision resulted in an increased incidence of fat necrosis.

  5. Brachytherapy for breast cancer

    International Nuclear Information System (INIS)

    Spikalovas, V.; Mudenas, A.; Karoesiene, E.; Mickevicius, R.

    1996-01-01

    In 1987-1995 a total of 347 patients with breast cancer underwent interstitial treatment. Two methods of irradiation were applied. 1. When patients refused surgery, external radiotherapy was given followed by implant radiotherapy for a dose of 20-30 Gy. Needle sources were applied for treatment with an increasing activity on the ends. The application of special template devices made it possible to implant radioactive sources in a strictly pre-set geometry. This allowed to place the sources in the necessary geometry for the whole course of irradiation. Dosimetric planning was performed in Gray-equivalents to a selected isodose curve mostly 85%. Treatment time was 20-50 hours. 2. In cases when the tumour was localized in the medial quadrant of breast, interstitial therapy was applied to the parasternal lymph nodes. During mastectomy catheters were placed in a. thoracica interna of the corresponding side. On the first or second postoperative day flexible radioactive sources were inserted into catheters. Their active length was 10-12 cm. Irradiation dose at a distance of 2 cm from the centre of source was 40-45 Gy. Results: There was minimum radiation effect on the adjusting organs and tissues. Local recurrence of tumour in the region of irradiation was in 6 patients. Conclusions: The application of interstitial radiotherapy in treatment of breast cancer is effective and the results of radiation treatment are encouraging

  6. The effect of interstitial 125I seeds implantation on intestinal wall: a pathological observation in experimental dogs

    International Nuclear Information System (INIS)

    Ning Houfa; Zhang Fenglian; Shen An; Cao Guiwen; Cui Xinjiang

    2010-01-01

    Objective: To observe the radiation injury of the bowel wall due to the implantation of interstitial 125 I seeds in experimental dogs. Methods: A total of 12 healthy male dogs were randomly and equally divided into 3 experimental groups and 1 control group, with 3 dogs in each group.In the experimental groups, two 125 I seeds with the active radiation dose of 0.8mCi were symmetrically implanted under the serous membrane of the dog's small intestinal wall. The dogs were fed for 14 days (group A), for one month (group B) and for two months (group C) respectively when the animals were scheduled to be sacrificed. The dogs' general condition was observed till they were sacrificed. The seed-implanting intestinal segments were then removed and dyed with HE staining method for electronic microscopic exam. The histopathologic findings were recorded and the results were compared between four groups. Results: No obvious histopathological changes were found in the dog's bowel wall 14 days after the implantation. One month after the procedure cellular injury was observed under electronic microscope, and two months after the operation partial fibrosis of the intestinal wall appeared but no ulceration or perforation occurred. Conclusion: The implantation of 125 I seeds can cause reversible cellular injuries of the intestinal wall in experimental dogs, the degree of the damage reaches its peak at one month after the implant when the partial fibrosis of bowel wall becomes evident. However, the seeds do not cause any serious complications, such as ulceration or perforation. (authors)

  7. CT-image based conformal high-dose rate brachytherapy boost in the conservative treatment of stage I - II breast cancer - introducing the procedure

    International Nuclear Information System (INIS)

    Kubaszewska, M.; Skowronek, J.; Chichel, A.; Kanikowski, M.; Dymnicka, M.

    2008-01-01

    Aim: Breast-conserving surgery (BCS) followed by radiotherapy (RT) has become the standard treatment for the majority of patients with early breast cancer. With regard to boost technique some disagreements are found between groups that are emphasizing the value of electron boost treatment and groups pointing out the value of interstitial brachytherapy (BT) boost treatment. We present the preliminary results in treating selected patients with early-stage breast cancer using high-dose-rate brachytherapy (HD R-BT) as a boost after breast conservation therapy (BCT). Materials/Methods: Between January 2006 and August 2007, a total of 58 female patients with first and second stage breast cancer underwent BCT. This therapeutic procedure involves BCS, whole breast radiation therapy (WBRT) and additional irradiation to the tumour bed (boost) using interstitial HDR-BT via flexible implant tubes. A 10 Gy boost dose was received by all patients. The treatment planning was based on CT-guided 3D (three-dimensional) reconstruction of the surgical clips, implant tubes and critical structures localization (skin and ribs). The accuracy of tumour bed localization, the conformity of planning target volume and treated volume were analyzed. Results: The evaluations of implant parameters involved the use of: dose volume histogram (DVH), the volume encompassed by the 100% reference isodose surface (V100%), the high dose volumecalculation (V150%, V200%, V300%), the dose non-uniformity ratio (DNR), and the conformity index (COIN). Our results were as follows: the mean PTV volume, the mean high dose volume (V150%; V200%; V300%), the DNR and COIN mean value were estimated at 57.38, 42.98, 21.38, 7.90, 0.52 and 0.83 respectively. Conclusions: CT-guided 3D HDR-BT is most appropriate for planning the boost procedure after BT especially in large breast volume, in cases with a deep seated tumour bed, as well as in patients with high risk for local recurrences. This technique reduces the

  8. Fully automated MRI-guided robotics for prostate brachytherapy

    International Nuclear Information System (INIS)

    Stoianovici, D.; Vigaru, B.; Petrisor, D.; Muntener, M.; Patriciu, A.; Song, D.

    2008-01-01

    The uncertainties encountered in the deployment of brachytherapy seeds are related to the commonly used ultrasound imager and the basic instrumentation used for the implant. An alternative solution is under development in which a fully automated robot is used to place the seeds according to the dosimetry plan under direct MRI-guidance. Incorporation of MRI-guidance creates potential for physiological and molecular image-guided therapies. Moreover, MRI-guided brachytherapy is also enabling for re-estimating dosimetry during the procedure, because with the MRI the seeds already implanted can be localised. An MRI compatible robot (MrBot) was developed. The robot is designed for transperineal percutaneous prostate interventions, and customised for fully automated MRI-guided brachytherapy. With different end-effectors, the robot applies to other image-guided interventions of the prostate. The robot is constructed of non-magnetic and dielectric materials and is electricity free using pneumatic actuation and optic sensing. A new motor (PneuStep) was purposely developed to set this robot in motion. The robot fits alongside the patient in closed-bore MRI scanners. It is able to stay fully operational during MR imaging without deteriorating the quality of the scan. In vitro, cadaver, and animal tests showed millimetre needle targeting accuracy, and very precise seed placement. The robot tested without any interference up to 7T. The robot is the first fully automated robot to function in MRI scanners. Its first application is MRI-guided seed brachytherapy. It is capable of automated, highly accurate needle placement. Extensive testing is in progress prior to clinical trials. Preliminary results show that the robot may become a useful image-guided intervention instrument. (author)

  9. Direction Modulated Brachytherapy for Treatment of Cervical Cancer. II: Comparative Planning Study With Intracavitary and Intracavitary–Interstitial Techniques

    Energy Technology Data Exchange (ETDEWEB)

    Han, Dae Yup [Department of Medical Physics, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of California, San Francisco, San Francisco, California (United States); Department of Electrical and Computer Engineering, University of California, San Diego, La Jolla, California (United States); Safigholi, Habib; Soliman, Abraam [Department of Medical Physics, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Physical Sciences, Sunnybrook Research Institute, Toronto, Ontario (Canada); Ravi, Ananth [Department of Medical Physics, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Leung, Eric [Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Department of Radiation Oncology, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Scanderbeg, Daniel J. [Department of Radiation Medicine and Applied Sciences, University of California, San Diego, La Jolla, California (United States); Liu, Zhaowei [Department of Electrical and Computer Engineering, University of California, San Diego, La Jolla, California (United States); Owrangi, Amir [Department of Medical Physics, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Song, William Y., E-mail: william.song@sunnybrook.ca [Department of Medical Physics, Sunnybrook Health Sciences Centre, Toronto, Ontario (Canada); Physical Sciences, Sunnybrook Research Institute, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada)

    2016-10-01

    Purpose: To perform a comprehensive comparative planning study evaluating the utility of the proposed direction modulated brachytherapy (DMBT) tandem applicator against standard applicators, in the setting of image guided adaptive brachytherapy of cervical cancer. Methods and Materials: A detailed conceptual article was published in 2014. The proposed DMBT tandem applicator has 6 peripheral grooves of 1.3-mm width, along a 5.4-mm-thick nonmagnetic tungsten alloy rod of density 18.0 g/cm{sup 3}, capable of generating directional dose profiles. We performed a comparative planning study with 45 cervical cancer patients enrolled consecutively in the prospective observational EMBRACE study. In all patients, MRI-based planning was performed while utilizing various tandem-ring (27 patients) and tandem-ring-needles (18 patients) applicators, in accordance with the Groupe Européen de Curiethérapie–European Society for Radiotherapy and Oncology recommendations. For unbiased comparisons, all cases were replanned with an in-house–developed inverse optimization code while enforcing a uniform set of constraints that are reflective of the clinical practice. All plans were normalized to the same high-risk clinical target volume D90 values achieved in the original clinical plans. Results: In general, if the standard tandem was replaced with the DMBT tandem while maintaining all other planning conditions the same, there was consistent improvement in the plan quality. For example, among the 18 tandem-ring-needles cases, the average D2cm{sup 3} reductions achieved were −2.48% ± 11.03%, −4.45% ± 5.24%, and −5.66% ± 6.43% for the bladder, rectum, and sigmoid, respectively. An opportunity may also exist in avoiding use of needles altogether for when the total number of needles required is small (approximately 2 to 3 needles or less), if DMBT tandem is used. Conclusions: Integrating the novel DMBT tandem onto both intracavitary and intracavitary–interstitial

  10. ALGEBRA: ALgorithm for the heterogeneous dosimetry based on GEANT4 for BRAchytherapy.

    Science.gov (United States)

    Afsharpour, H; Landry, G; D'Amours, M; Enger, S; Reniers, B; Poon, E; Carrier, J-F; Verhaegen, F; Beaulieu, L

    2012-06-07

    Task group 43 (TG43)-based dosimetry algorithms are efficient for brachytherapy dose calculation in water. However, human tissues have chemical compositions and densities different than water. Moreover, the mutual shielding effect of seeds on each other (interseed attenuation) is neglected in the TG43-based dosimetry platforms. The scientific community has expressed the need for an accurate dosimetry platform in brachytherapy. The purpose of this paper is to present ALGEBRA, a Monte Carlo platform for dosimetry in brachytherapy which is sufficiently fast and accurate for clinical and research purposes. ALGEBRA is based on the GEANT4 Monte Carlo code and is capable of handling the DICOM RT standard to recreate a virtual model of the treated site. Here, the performance of ALGEBRA is presented for the special case of LDR brachytherapy in permanent prostate and breast seed implants. However, the algorithm is also capable of handling other treatments such as HDR brachytherapy.

  11. A comprehensive review of prostate cancer brachytherapy: defining an optimal technique

    International Nuclear Information System (INIS)

    Vicini, Frank A.; Kini, Vijay R.; Edmundson, Gregory B.S.; Gustafson, Gary S.; Stromberg, Jannifer; Martinez, Alvaro

    1999-01-01

    Purpose: A comprehensive review of prostate cancer brachytherapy literature was performed to determine if an optimal method of implantation could be identified, and to compare and contrast techniques currently in use. Methods and Materials: A MEDLINE search was conducted to obtain all articles in the English language on prostate cancer brachytherapy from 1985 through 1998. Articles were reviewed and grouped to determine the primary technique of implantation, the method or philosophy of source placement and/or dose specification, the technique to evaluate implant quality, overall treatment results (based upon pretreatment prostate specific antigen, (PSA), and biochemical control) and clinical, pathological or biochemical outcome based upon implant quality. Results: A total of 178 articles were identified in the MEDLINE database. Of these, 53 studies discussed evaluable techniques of implantation and were used for this analysis. Of these studies, 52% used preoperative ultrasound to determine the target volume to be implanted, 16% used preoperative computerized tomography (CT) scans, and 18% placed seeds with an open surgical technique. An additional 11% of studies placed seeds or needles under ultrasound guidance using interactive real-time dosimetry. The number and distribution of radioactive sources to be implanted or the method used to prescribe dose was determined using nomograms in 27% of studies, a least squares optimization technique in 11%, or not stated in 35%. In the remaining 26%, sources were described as either uniformly, differentially, or peripherally placed in the gland. To evaluate implant quality, 28% of studies calculated some type of dose-volume histogram, 21% calculated the matched peripheral dose, 19% the minimum peripheral dose, 14% used some type of CT-based qualitative review and, in 18% of studies, no implant quality evaluation was mentioned. Six studies correlated outcome with implant dose. One study showed an association of implant dose

  12. Third-party brachytherapy source calibrations and physicist responsibilities: Report of the AAPM Low Energy Brachytherapy Source Calibration Working Group

    International Nuclear Information System (INIS)

    Butler, Wayne M.; Bice, William S. Jr.; DeWerd, Larry A.; Hevezi, James M.; Huq, M. Saiful; Ibbott, Geoffrey S.; Palta, Jatinder R.; Rivard, Mark J.; Seuntjens, Jan P.; Thomadsen, Bruce R.

    2008-01-01

    The AAPM Low Energy Brachytherapy Source Calibration Working Group was formed to investigate and recommend quality control and quality assurance procedures for brachytherapy sources prior to clinical use. Compiling and clarifying recommendations established by previous AAPM Task Groups 40, 56, and 64 were among the working group's charges, which also included the role of third-party handlers to perform loading and assay of sources. This document presents the findings of the working group on the responsibilities of the institutional medical physicist and a clarification of the existing AAPM recommendations in the assay of brachytherapy sources. Responsibility for the performance and attestation of source assays rests with the institutional medical physicist, who must use calibration equipment appropriate for each source type used at the institution. Such equipment and calibration procedures shall ensure secondary traceability to a national standard. For each multi-source implant, 10% of the sources or ten sources, whichever is greater, are to be assayed. Procedures for presterilized source packaging are outlined. The mean source strength of the assayed sources must agree with the manufacturer's stated strength to within 3%, or action must be taken to resolve the difference. Third party assays do not absolve the institutional physicist from the responsibility to perform the institutional measurement and attest to the strength of the implanted sources. The AAPM leaves it to the discretion of the institutional medical physicist whether the manufacturer's or institutional physicist's measured value should be used in performing dosimetry calculations

  13. Monte Carlo simulations and radiation dosimetry measurements of 142Pr capillary tube-based radioactive implant (CTRI). A new structure for brachytherapy sources

    International Nuclear Information System (INIS)

    Bakht, M.K.; Haddadi, A.; Sadeghi, M.; Ahmadi, S.J.; Sadjadi, S.S.; Tenreiro, C.

    2013-01-01

    Previously, a promising β - -emitting praseodymium-142 glass seed was proposed for brachytherapy of prostate cancer. In accordance with the previous study, a 142 Pr capillary tube-based radioactive implant (CTRI) was suggested as a source with a new structure to enhance application of β - -emitting radioisotopes such as 142 Pr in brachytherapy. Praseodymium oxide powder was encapsulated in a glass capillary tube. Then, a thin and flexible fluorinated ethylene propylene Teflon layer sealed the capillary tube. The source was activated in the Tehran Research Reactor by the 141 Pr(n, γ) 142 Pr reaction. Measurements of the dosimetric parameters were performed using GafChromic radiochromic film. In addition, the dose rate distribution of 142 Pr CTRI was calculated by modeling 142 Pr source in a water phantom using Monte Carlo N-Particle Transport (MCNP5) Code. The active source was unreactive and did not leak in water. In comparison with the earlier proposed 142 Pr seed, the suggested source showed similar desirable dosimetric characteristics. Moreover, the 142 Pr CTRI production procedure may be technically and economically more feasible. The mass of praseodymium in CTRI structure could be greater than that of the 142 Pr glass seed; therefore, the required irradiation time and the neutron flux could be reduced. A 142 Pr CTRI was proposed for brachytherapy of prostate cancer. The dosimetric calculations by the experimental measurements and Monte Carlo simulation were performed to fulfill the requirements according to the American Association of Physicists in Medicine recommendations before the clinical use of new brachytherapy sources. The characteristics of the suggested source were compared with those of the previously proposed 142 Pr glass seed. (author)

  14. Development of virtual patient models for permanent implant brachytherapy Monte Carlo dose calculations: interdependence of CT image artifact mitigation and tissue assignment.

    Science.gov (United States)

    Miksys, N; Xu, C; Beaulieu, L; Thomson, R M

    2015-08-07

    This work investigates and compares CT image metallic artifact reduction (MAR) methods and tissue assignment schemes (TAS) for the development of virtual patient models for permanent implant brachytherapy Monte Carlo (MC) dose calculations. Four MAR techniques are investigated to mitigate seed artifacts from post-implant CT images of a homogeneous phantom and eight prostate patients: a raw sinogram approach using the original CT scanner data and three methods (simple threshold replacement (STR), 3D median filter, and virtual sinogram) requiring only the reconstructed CT image. Virtual patient models are developed using six TAS ranging from the AAPM-ESTRO-ABG TG-186 basic approach of assigning uniform density tissues (resulting in a model not dependent on MAR) to more complex models assigning prostate, calcification, and mixtures of prostate and calcification using CT-derived densities. The EGSnrc user-code BrachyDose is employed to calculate dose distributions. All four MAR methods eliminate bright seed spot artifacts, and the image-based methods provide comparable mitigation of artifacts compared with the raw sinogram approach. However, each MAR technique has limitations: STR is unable to mitigate low CT number artifacts, the median filter blurs the image which challenges the preservation of tissue heterogeneities, and both sinogram approaches introduce new streaks. Large local dose differences are generally due to differences in voxel tissue-type rather than mass density. The largest differences in target dose metrics (D90, V100, V150), over 50% lower compared to the other models, are when uncorrected CT images are used with TAS that consider calcifications. Metrics found using models which include calcifications are generally a few percent lower than prostate-only models. Generally, metrics from any MAR method and any TAS which considers calcifications agree within 6%. Overall, the studied MAR methods and TAS show promise for further retrospective MC dose

  15. Sci-Thur PM – Brachytherapy 02: Positional accuracy in Pd-103 permanent breast seed implant (PBSI) brachytherapy at the Tom Baker Cancer Centre (TBCC)

    Energy Technology Data Exchange (ETDEWEB)

    Frederick, Amy; Watt, Elizabeth; Peacock, Michael; Husain, Siraj; Meyer, Tyler; Roumeliotis, Michael [University of Calgary, Tom Baker Cancer Centre (Canada)

    2016-08-15

    Purpose: This retrospective study aims to quantify the positional accuracy of seed delivery in permanent breast seed implant (PBSI) brachytherapy at the Tom Baker Cancer Centre (TBCC). Methods: Treatment planning and post-implant CT scans for 5 patients were rigidly registered using the MIM Symphony™ software (MIM Software, Cleveland, OH) and used to evaluate differences between planned and implanted seed positions. Total and directional seed displacements were calculated for each patient in a clinically relevant ‘needle coordinate system’, defined relative to the angle of fiducial needle insertion. Results: The overall average total seed displacement was 10±8 mm. Systematic seed displacements were observed in individual patients and the magnitude and direction of these offsets varied among patients. One patient showed a significant directional seed displacement in the shallow-deep direction compared with the other four patients. With the exception of this one patient outlier, no significant systematic directional displacements in the needle coordinate system were observed for this cohort; the average directional displacements were −1±5 mm, 2±3 mm, and −2±4 mm in the shallow-deep, up-down, and right-left directions respectively. Conclusion: With the exception of one patient outlier, the magnitude of seed displacements were relatively consistent among patients. The results indicate that the shallow-deep direction possesses the largest uncertainty for the seed delivery method used at the TBCC. The relatively large uncertainty in seed placement in this direction is expected, as this is the direction of needle insertion. Further work will involve evaluating deflections of delivered needle tracks from their planned positions.

  16. Clinical implementation of a quality assurance program in HDR brachytherapy by in vivo dosimetry with diodes

    International Nuclear Information System (INIS)

    Alecu, R.; Feldmeier, J.J.; Court, W.S.; Alecu, M.; Orton, C.G.

    1996-01-01

    Despite the possibilities of in vivo dosimetry with diodes (e.g. control of dose to organs at risk, recorded confirmational measurements of the dose actually delivered, check of the whole treatment chain, avoidance of misadministrations, etc.) and the fact that it has proven to be very useful as part of a departmental QA program in external beam therapy, few attempts to implement it for HDR brachytherapy procedures have been reported. The reason for this is probably that there are significant technical challenges that must be met prior to its clinical use. The purpose of this study is to investigate the practicability and usefulness of dose measurements for brachytherapy patients in daily clinical practice. In our clinic a high precision patient dosimetry method has been developed, based on the use of silicon diodes. First, calibration factors have been determined under 'reference' irradiation conditions. Secondly, correction factors have been evaluated for situations deviating from the reference conditions, i.e. for different distances from the implanted sources, tissue heterogeneities, presence of different type of applicators, etc. For certain intracavitary, interstitial and surface mold applications this procedure has proven to be sufficiently accurate to allow dose determinations with diodes to be in good agreement with the expected values, i.e. calculated by the treatment planning system (VariSource unit) and checked by ion chamber measurements. The results of in vivo measurements are discussed along with the possibilities and limitations of the employed techniques

  17. Role of TPS in 125I brachytherapy for orbital tumors

    International Nuclear Information System (INIS)

    Ren Ling; Dai Haojie; Li Quan

    2012-01-01

    Objective: To investigate the role of TPS in 125 I brachytherapy for orbital tumors. Methods: Sixty-six patients with orbital tumor treated with 125 I seeds from 2005 to 2009 were retrospectively analyzed. Forty-three patients were treated using TPS guided brachytherapy and the prescribed dose was 140 Gy. Other 23 patients were treated without TPS but simply implanted with 125 I seeds at 1 cm intervals in parallel with each other intraoperatively. CT and TPS quality verification were performed postoperatively in all patients. Also, CT and (or) MRI examination were performed at 3, 6, 12 and 24 months after brachytherapy for follow-up. χ 2 test and Kaplan-Meier survival analysis with log-rank significance test were used with SPSS 17.0. Results: A total of 1070 125 I seeds were implanted in 66 cases, on average, (16.2 ± 7.3) seeds for each patient. The satisfaction rates of postoperative quality verification in patients with and without TPS pre-plans were 79.07% (34/43) and 43.48% (10/23) respectively (χ 2 =8.542, P=0.003). Ten patients were lost in follow-up. Local recurrence rates in patients with favorable postoperative quality verification were 0 (0/37) in 3 months, 6.25% (2/32) in 6 months, 13.64% (3/22) in 12 months and 3/9 in 24 months respectively, which were significantly different from those (5.26% (1/19), 16.67% (3/18), 30.77% (4/13), 6/6) in the patients with inferior postoperative quality verification (χ 2 =9.017, P=0.0003). Conclusions: TPS plays an important role in 125 I brachytherapy for orbital tumors. Also, postoperative quality verification by TPS may help predict the local recurrence after brachytherapy. (authors)

  18. Intraoperative real-time planned conformal prostate brachytherapy: Post-implantation dosimetric outcome and clinical implications

    International Nuclear Information System (INIS)

    Zelefsky, Michael J.; Yamada, Yoshiya; Cohen, Gil'ad N.; Sharma, Neha; Shippy, Alison M.; Fridman, David; Zaider, Marco

    2007-01-01

    Purpose: To report the dosimetric outcome of patients with clinically localized prostate cancer treated with I-125 permanent implantation using an intraoperative real-time conformal planning technique. Methods and materials: Five hundred and sixty-two patients with prostate cancer were treated with I-125 permanent interstitial implantation using a transrectal ultrasound-guided approach. Real-time intraoperative treatment planning software that incorporates inverse planning optimization was used. Dose-volume constraints for this inverse-planning system included: prostate V100 ≥95%, maximal urethral dose ≤120%, and average rectal dose 3 of the rectum was exposed to the prescription dose, the incidence of late grade 2 toxicity rectal toxicity was 9% compared to 4% for smaller volumes of the rectum exposed to similar doses (p = 0.003). No dosimetric parameter in these patients with tight dose confines for the urethra influenced acute or late urinary toxicity. Conclusion: Real-time intraoperative planning was associated with a 90% consistency of achieving the planned intraoperative dose constraints for target coverage and maintaining planned urethral and rectal constraints in a high percentage of implants. Rectal volumes of ≥2.5 cm 3 exposed to the prescription doses were associated with an increased incidence of grade 2 rectal bleeding. Further enhancements in imaging guidance for optimal seed deposition are needed to guarantee optimal dose distribution for all patients. Whether such improvements lead to further reduction in acute and late morbidities associated with therapy requires further study

  19. Evaluation of two intracavitary high-dose-rate brachytherapy devices for irradiating additional and irregularly shaped volumes of breast tissue

    International Nuclear Information System (INIS)

    Lu, Sharon M.; Scanderbeg, Daniel J.; Barna, Patrick; Yashar, William; Yashar, Catheryn

    2012-01-01

    The SAVI and Contura breast brachytherapy applicators represent 2 recent advancements in brachytherapy technology that have expanded the number of women eligible for accelerated partial breast irradiation in the treatment of early-stage breast cancer. Early clinical experience with these 2 single-entry, multichannel high-dose-rate brachytherapy devices confirms their ease of use and dosimetric versatility. However, current clinical guidelines for SAVI and Contura brachytherapy may result in a smaller or less optimal volume of treated tissue compared with traditional interstitial brachytherapy. This study evaluates the feasibility of using the SAVI and Contura to irradiate larger and irregularly shaped target volumes, approaching what is treatable with the interstitial technique. To investigate whether additional tissue can be treated, 17 patients treated with the SAVI and 3 with the Contura were selected. For each patient, the planning target volume (PTV) was modified to extend 1.1 cm, 1.3 cm, and 1.5 cm beyond the tumor bed cavity. To evaluate dose conformance to an irregularly shaped target volume, 9 patients treated with the SAVI and 3 with the Contura were selected from the original 20 patients. The following asymmetric PTV margin combinations were assessed for each patient: 1.5/0.3, 1.3/0.3, and 1.1/0.3 cm. For all patients, treatment planning was performed, adopting the National Surgical Adjuvant Breast and Bowel Project guidelines, and dosimetric comparisons were made. The 6–1 and 8–1 SAVI devices can theoretically treat a maximal tissue margin of 1.5 cm and an asymmetric PTV with margins ranging from 0.3 to 1.5 cm. The 10–1 SAVI and Contura can treat a maximal margin of 1.3 cm and 1.1 cm, respectively, and asymmetric PTV with margins ranging from 0.3–1.3 cm. Compared with the Contura, the SAVI demonstrated greater dosimetric flexibility. Risk of developing excessive hot spots increased with the size of the SAVI device. Both the SAVI and Contura

  20. Equivalence of hyperfractionated and continuous brachytherapy in a rat tumor model and remarkable effectiveness when preceded by external irradiation

    International Nuclear Information System (INIS)

    Veninga, Theo; Visser, Andries G.; Berg, Ad P. van den; Hooije, Christel van; Geel, Cornelis A.J.F. van; Levendag, Peter C.

    2001-01-01

    Purpose: In clinical brachytherapy, there is a tendency to replace continuous low-dose-rate (LDR) irradiation by either single-dose or fractionated high-dose-rate (HDR) irradiation. In this study, the equivalence of LDR treatments and fractionated HDR (2 fractions/day) or pulsed-dose-rate (PDR, 4 fractions/day) schedules in terms of tumor cure was investigated in an experimental tumor model. Methods and Materials: Tumors (rat rhabdomyosarcoma R1M) were grown s.c. in the flank of rats and implanted with 4 catheters guided by a template. All interstitial radiation treatment (IRT) schedules were given in the same geometry. HDR was given using an 192 Ir single-stepping source. To investigate small fraction sizes, part of the fractionated HDR and PDR schedules were applied after an external irradiation (ERT) top-up dose. The endpoint was the probability of tumor control at 150 days after treatment. Cell survival was estimated by excision assay. Results: Although there was no fractionation effect for fractionated HDR given in 1 or 2 fractions per day, TCD 50 -values were substantially lower than that for LDR. A PDR schedule with an interfraction interval of 3 h (4 fractions/day), however, was equivalent to LDR. The combination of ERT and IRT resulted in a remarkably increased tumor control probability in all top-up regimens, but no difference was found between 2 or 4 fractions/day. Catheter implantation alone decreased the TCD 50 for single-dose ERT already by 17.4 Gy. Cell viability assessed at 24 h after treatment demonstrated an increased effectiveness of interstitial treatment, but, after 10 Gy ERT followed by 10 Gy IRT (24-h interval), it was not less than that calculated for the combined effect of these treatments given separately. Conclusion: In full fractionation schedules employing large fractions and long intervals, the sparing effect of sublethal damage repair may be significantly counteracted by reoxygenation. During 3-h intervals, however, repair may be

  1. Brachytherapy of penis cancer: experience of the Gustave-Roussy Institute; Curietherapie des cancers de verge: experience de l'Institut Gustave-Roussy

    Energy Technology Data Exchange (ETDEWEB)

    Crevoisier, R. de; Wibault, P.; M' barek, B.; Gerbaulet, A.; Haie-Meder, C. [Institut Gustave-Roussy, 94 - Villejuif (France)

    2006-11-15

    The interstitial brachytherapy is a conservative treatment of epidermoid carcinomas of the penis under 4 centimeters and does not invade the erectile tissue. the great majority of recurrences is under control by a surgery treatment. (N.C.)

  2. Erectile function after prostate brachytherapy

    International Nuclear Information System (INIS)

    Merrick, Gregory S.; Butler, Wayne M.; Wallner, Kent E.; Galbreath, Robert W.; Anderson, Richard L.; Kurko, Brian S.; Lief, Jonathan H.; Allen, Zachariah A.

    2005-01-01

    Purpose: To evaluate erectile function after permanent prostate brachytherapy using a validated patient-administered questionnaire and to determine the effect of multiple clinical, treatment, and dosimetric parameters on penile erectile function. Methods and materials: A total of 226 patients with preimplant erectile function determined by the International Index of Erectile Function (IIEF) questionnaire underwent permanent prostate brachytherapy in two prospective randomized trials between February 2001 and January 2003 for clinical Stage T1c-T2c (2002 American Joint Committee on Cancer) prostate cancer. Of the 226 patients, 132 were potent before treatment and, of those, 128 (97%) completed and returned the IIEF questionnaire after brachytherapy. The median follow-up was 29.1 months. Potency was defined as an IIEF score of ≥13. The clinical, treatment, and dosimetric parameters evaluated included patient age; preimplant IIEF score; clinical T stage; pretreatment prostate-specific antigen level; Gleason score; elapsed time after implantation; preimplant nocturnal erections; body mass index; presence of hypertension or diabetes mellitus; tobacco consumption; the volume of the prostate gland receiving 100%, 150%, and 200% of the prescribed dose (V 100/150/200 ); the dose delivered to 90% of the prostate gland (D 90 ); androgen deprivation therapy; supplemental external beam radiotherapy (EBRT); isotope; prostate volume; planning volume; and radiation dose to the proximal penis. Results: The 3-year actuarial rate of potency preservation was 50.5%. For patients who maintained adequate posttreatment erectile function, the preimplant IIEF score was 29, and in patients with brachytherapy-related ED, the preimplant IIEF score was 25. The median time to the onset of ED was 5.4 months. After brachytherapy, the median IIEF score was 20 in potent patients and 3 in impotent patients. On univariate analysis, the preimplant IIEF score, patient age, presence of nocturnal

  3. Evaluation of resins for use in brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Carvalho, Luiz Claudio F.M. Garcia; Ferraz, Wilmar Barbosa; Chrcanovic, Bruno Ramos; Santos, Ana Maria M., E-mail: ferrazw@cdtn.b, E-mail: amms@cdtn.b [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2011-07-01

    Brachytherapy is an advanced cancer treatment where radioactive seeds or sources are placed near or directly into the tumor thus reducing the radiation exposure in the surrounding healthy tissues. Prostate cancer can be treated with interstitial brachytherapy in initial stage of the disease in which tiny radioactive seeds with cylindrical geometry are used. Several kinds of seeds have been developed in order to obtain a better dose distribution around them and with a lower cost manufacturing. These seeds consist of an encapsulation, a radionuclide carrier, and X-ray marker. Among the materials that have potential for innovation in the construction of seeds, biocompatible resins appear as an important option. In this paper, we present some characterization results with Fourier transform infrared spectroscopic (FTIR) and ultraviolet-visible spectroscopy (UV-vis) performed on two types of resins in which curing temperatures for each one were varied as also the results of coatings with these resins under titanium substrates. Interactions of these resins in contact with the simulated body fluid were evaluated by atomic force microscopy, scanning electron microscopy, and energy dispersive X-ray spectroscopy. (author)

  4. Brachytherapy-related complications for medically inoperable Stage I endometrial carcinoma

    International Nuclear Information System (INIS)

    Chao, Clifford K. S.; Grigsby, Perry W.; Perez, Carlos A.; Camel, H. Marvin; Kao, Ming-Shian; Galakatos, Andrew E.; Boyle, Walter A.

    1995-01-01

    Purpose: The current study was conducted to investigate the incidence and risk factors for medical complications associated with low dose rate brachytherapy in patients with medically inoperable Stage I endometrial cancer treated with irradiation alone. Methods and Materials: From 1965 through 1991 at Mallinckrodt Institute of Radiology, 150 implants were performed on 96 patients who were deemed medically unfit for hysterectomy because of advanced age, obesity, and various medical problems. The records of these patients were examined retrospectively to determine the incidence of medical complications that occurred in the first 30 days following the initiation of brachytherapy. The association of risk factors that precluded major surgery and the occurrence of brachytherapy-related complications was examined by logistic regression. Results: Of these 96 patients, 40 patients were older than 75 years, and 31 patients were deemed morbidly obese. Medical problems included hypertension in 45 patients, and diabetes in 37; there was a history of congestive heart failure in 23, stroke in 11, myocardial infarction in 10, and thromboembolism in 8. There were concurrent malignancies in five patients. Implants were performed using intrauterine Simon-Heyman capsules, tandems, and vaginal ovoids in all patients. General anesthesia was used for 98 implants, spinal anesthesia for 26, local anesthesia for 25, and epidural anesthesia for 1. The duration of anesthesia ranged from 30 to 120 min (median, 60 min). The duration of radioisotope application ranged from 11 to 96 h (median, 46 h). Preventive measures included low dose subcutaneous heparin in 55 patients (since 1978), and intermittent pneumatic compression boots in 29 (since 1985). Four patients developed life-threatening complications including myocardial infarction (two patients), congestive heart failure (one patient), and pulmonary embolism (one patient). Two of these four patients died; one with a myocardial infarction and

  5. Comparison of Combined X-Ray Radiography and Magnetic Resonance (XMR) Imaging-Versus Computed Tomography-Based Dosimetry for the Evaluation of Permanent Prostate Brachytherapy Implants

    International Nuclear Information System (INIS)

    Acher, Peter; Rhode, Kawal; Morris, Stephen; Gaya, Andrew; Miquel, Marc; Popert, Rick; Tham, Ivan; Nichol, Janette; McLeish, Kate; Deehan, Charles; Dasgupta, Prokar; Beaney, Ronald; Keevil, Stephen F.

    2008-01-01

    Purpose: To present a method for the dosimetric analysis of permanent prostate brachytherapy implants using a combination of stereoscopic X-ray radiography and magnetic resonance (MR) imaging (XMR) in an XMR facility, and to compare the clinical results between XMR- and computed tomography (CT)-based dosimetry. Methods and Materials: Patients who had received nonstranded iodine-125 permanent prostate brachytherapy implants underwent XMR and CT imaging 4 weeks later. Four observers outlined the prostate gland on both sets of images. Dose-volume histograms (DVHs) were derived, and agreement was compared among the observers and between the modalities. Results: A total of 30 patients were evaluated. Inherent XMR registration based on prior calibration and optical tracking required a further automatic seed registration step that revealed a median root mean square registration error of 4.2 mm (range, 1.6-11.4). The observers agreed significantly more closely on prostate base and apex positions as well as outlining contours on the MR images than on those from CT. Coefficients of variation were significantly higher for observed prostate volumes, D90, and V100 parameters on CT-based dosimetry as opposed to XMR. The XMR-based dosimetry showed little agreement with that from CT for all observers, with D90 95% limits of agreement ranges of 65, 118, 79, and 73 Gy for Observers 1, 2, 3, and 4, respectively. Conclusions: The study results showed that XMR-based dosimetry offers an alternative to other imaging modalities and registration methods with the advantages of MR-based prostate delineation and confident three-dimensional reconstruction of the implant. The XMR-derived dose-volume histograms differ from the CT-derived values and demonstrate less interobserver variability

  6. Physics and quality assurance for high dose rate brachytherapy

    International Nuclear Information System (INIS)

    Anderson, Lowell L.

    1995-01-01

    Purpose: To review the physical aspects of high dose rate (HDR) brachytherapy, including commissioning and quality assurance, source calibration and dose distribution measurements, and treatment planning methods. Following the introduction of afterloading in brachytherapy, development efforts to make it 'remote' culminated in 1964 with the near-simultaneous appearance of remote afterloaders in five major medical centers. Four of these machines were 'high dose rate', three employing 60Co and one (the GammaMed) using a single, cable-mounted 192Ir source. Stepping-motor source control was added to the GammaMed in 1974, making it the precursor of modern remote afterloaders, which are now suitable for interstitial as well as intracavitary brachytherapy by virtue of small source-diameter and indexer-accessed multiple channels. Because the 192Ir sources currently used in HDR remote afterloaders are supplied at a nominal air-kerma strength of 11.4 cGy cm2 s-1 (10 Ci), are not collimated in clinical use, and emit a significant fraction (15%) of photons at energies greater than 600 keV, shielding and facility design must be undertaken as carefully and thoroughly as for external beam installations. Licensing requirements of regulatory agencies must be met with respect both to maximum permissible dose limits and to the existence and functionality of safety devices (door interlocks, radiation monitors, etc.). Commissioning and quality assurance procedures that must be documented for HDR remote afterloading relate to (1) machine, applicator, guide-tube, and facility functionality checks, (2) source calibration, (3) emergency response readiness, (4) planning software evaluation, and (5) independent checks of clinical dose calculations. Source calibration checks must be performed locally, either by in-air measurement of air kerma strength or with a well ionization chamber calibrated (by an accredited standards laboratory) against an in-air measurement of air kerma strength for the

  7. Silicone implant incompatibility syndrome (SIIS) in a 57-year-old woman with unilateral silicone breast implant

    DEFF Research Database (Denmark)

    Schierbeck, Juliane; Davidsen, Jesper Rømhild; Grindsted Nielsen, Sanne

    2017-01-01

    implants can lead to different interstitial lung manifestations predominantly with granuloma evolvement, leading to the so-called silicone implant incompatibility syndrome (SIIS). This case describes a 57-year-old woman with multiple lung infiltrations and a left-sided breast implant. The implant had been...

  8. He implantation induced nanovoids in crystalline Si

    International Nuclear Information System (INIS)

    Kilpelaeinen, S.; Kuitunen, K.; Slotte, J.; Tuomisto, F.; Bruno, E.; Mirabella, S.; Priolo, F.

    2009-01-01

    Positron annihilation spectroscopy (PAS) in Doppler broadening mode was used to study the vacancy profile of crystalline Si after He and B implantation and subsequent annealing. In the He-implanted samples two different void layers were observed, one consisting of large voids at the projected range of He and another containing 'nanovoids' slightly larger than divacancies at roughly halfway between R p of He and the surface. The nanovoid layer was shown to be absent from samples co-implanted with B, implying that interstitials created during B implantation get trapped in the nanovoids and fill them, thus hindering interstitial-mediated B diffusion.

  9. Isotope selection for patients undergoing prostate brachytherapy

    International Nuclear Information System (INIS)

    Cha, Christine M.; Potters, Louis; Ashley, Richard; Freeman, Katherine; Wang Xiaohong; Waldbaum, Robert; Leibel, Steven

    1999-01-01

    Purpose: Ultrasound-guided trans perineal interstitial permanent prostate brachytherapy (TIPPB) is generally performed with either 103 Pd or 125 I. The use of 125 I for low Gleason score tumors and 103 Pd for higher Gleason scores has been suggested based on isotope dose rate and cell doubling time observed in in vitro studies. While many centers follow these isotope selection criteria, other centers have elected to use only a single isotope, regardless of Gleason score. No clinical data have been published comparing these isotopes. This study was undertaken to compare outcomes between 125 I and 103 Pd in a matched pair analysis for patients undergoing prostate brachytherapy. Methods and Materials: Six hundred forty-eight consecutively treated patients with clinically confined prostate cancer underwent TIPPB between June 1992 and February 1997. Five hundred thirty-two patients underwent TIPPB alone, whereas 116 received pelvic external beam irradiation and TIPPB. Ninety-three patients received androgen deprivation therapy prior to TIPPB. The prescribed doses for TIPPB were 160 Gy for 125 I (pre-TG43) and 120 Gy for 103 Pd. Patients treated with combination therapy received 41.4 or 45 Gy (1.8 Gy/fraction) external beam irradiation followed by a 3- to 5-week break and then received either a 120-Gy 125 I or a 90-Gy 103 Pd implant. Until November 1994, all patients underwent an 125 I implant after which the isotope selection was based on either Gleason score (Gleason score 2-5: 125 I; Gleason 5-8: 103 Pd) or isotope availability. A matched pair analysis was performed to assess any difference between isotopes. Two hundred twenty-two patients were matched according to Gleason score, prostate-specific antigen (PSA), and stage. PSA relapse-free survival (PSA-RFS) was calculated based on the American Society for Therapeutic Radiology and Oncology (ASTRO) Consensus Group definition of failure. Kaplan-Meier actuarial survival curves were compared to assess differences in

  10. The use of nomograms in LDR-HDR prostate brachytherapy.

    Science.gov (United States)

    Pujades, Ma Carmen; Camacho, Cristina; Perez-Calatayud, Jose; Richart, José; Gimeno, Jose; Lliso, Françoise; Carmona, Vicente; Ballester, Facundo; Crispín, Vicente; Rodríguez, Silvia; Tormo, Alejandro

    2011-09-01

    The common use of nomograms in Low Dose Rate (LDR) permanent prostate brachytherapy (BT) allows to estimate the number of seeds required for an implant. Independent dosimetry verification is recommended for each clinical dosimetry in BT. Also, nomograms can be useful for dose calculation quality assurance and they could be adapted to High Dose Rate (HDR). This work sets nomograms for LDR and HDR prostate-BT implants, which are applied to three different institutions that use different implant techniques. Patients treated throughout 2010 till April 2011 were considered for this study. This example was chosen to be the representative of the latest implant techniques and to ensure consistency in the planning. A sufficient number of cases for both BT modalities, prescription dose and different work methodology (depending on the institution) were taken into account. The specific nomograms were built using the correlation between the prostate volume and some characteristic parameters of each BT modality, such as the source Air Kerma Strength, number of implanted seeds in LDR or total radiation time in HDR. For each institution and BT modality, nomograms normalized to the prescribed dose were obtained and fitted to a linear function. The parameters of the adjustment show a good agreement between data and the fitting. It should be noted that for each institution these linear function parameters are different, indicating that each centre should construct its own nomograms. Nomograms for LDR and HDR prostate brachytherapy are simple quality assurance tools, specific for each institution. Nevertheless, their use should be complementary to the necessary independent verification.

  11. The use of nomograms in LDR-HDR prostate brachytherapy

    Directory of Open Access Journals (Sweden)

    Ma Carmen Pujades

    2011-09-01

    Full Text Available Purpose: The common use of nomograms in Low Dose Rate (LDR permanent prostate brachytherapy (BT allowsto estimate the number of seeds required for an implant. Independent dosimetry verification is recommended for eachclinical dosimetry in BT. Also, nomograms can be useful for dose calculation quality assurance and they could be adaptedto High Dose Rate (HDR. This work sets nomograms for LDR and HDR prostate-BT implants, which are applied tothree different institutions that use different implant techniques. Material and methods: Patients treated throughout 2010 till April 2011 were considered for this study. This examplewas chosen to be the representative of the latest implant techniques and to ensure consistency in the planning. A sufficientnumber of cases for both BT modalities, prescription dose and different work methodology (depending on theinstitution were taken into account. The specific nomograms were built using the correlation between the prostatevo lume and some characteristic parameters of each BT modality, such as the source Air Kerma Strength, numberof implanted seeds in LDR or total radiation time in HDR. Results: For each institution and BT modality, nomograms normalized to the prescribed dose were obtained andfitted to a linear function. The parameters of the adjustment show a good agreement between data and the fitting.It should be noted that for each institution these linear function parameters are different, indicating that each centreshould construct its own nomograms. Conclusions: Nomograms for LDR and HDR prostate brachytherapy are simple quality assurance tools, specific foreach institution. Nevertheless, their use should be complementary to the necessary independent verification.

  12. A Dosimetric Comparison of Accelerated Partial Breast Irradiation Techniques: Multicatheter Interstitial Brachytherapy, Three-Dimensional Conformal Radiotherapy, and Supine Versus Prone Helical Tomotherapy

    International Nuclear Information System (INIS)

    Patel, Rakesh R.; Becker, Stewart J.; Das, Rupak K.; Mackie, Thomas R.

    2007-01-01

    Purpose: To compare dosimetrically four different techniques of accelerated partial breast irradiation (APBI) in the same patient. Methods and Materials: Thirteen post-lumpectomy interstitial brachytherapy (IB) patients underwent imaging with preimplant computed tomography (CT) in the prone and supine position. These CT scans were then used to generate three-dimensional conformal radiotherapy (3D-CRT) and prone and supine helical tomotherapy (PT and ST, respectively) APBI plans and compared with the treated IB plans. Dose-volume histogram analysis and the mean dose (NTD mean ) values were compared. Results: Planning target volume coverage was excellent for all methods. Statistical significance was considered to be a p value mean dose of 1.3 Gy 3 and 1.2 Gy 3 , respectively. Both of these methods were statistically significantly lower than the supine external beam techniques. Overall, all four methods yielded similar low doses to the heart. Conclusions: The use of IB and PT resulted in greater normal tissue sparing (especially ipsilateral breast and lung) than the use of supine external beam techniques of 3D-CRT or ST. However, the choice of APBI technique must be tailored to the patient's anatomy, lumpectomy cavity location, and overall treatment goals

  13. BED-Volume histograms calculation for routine clinical dosimetry in brachytherapy

    International Nuclear Information System (INIS)

    Galelli, M.; Feroldi, P.

    1995-01-01

    The consideration of volumes is essential in Brachytherapy clinical dosimetry (I.C.R.U). Indeed, several indices, all based on dose-volume histograms (DVHs), have been designed in order to evaluate: before the therapy the volumetric quality of different possible implant geometries; during the therapy the consistency of the real and the previsional implants. Radiobiological evaluations, considering the dose deposition temporal pattern of treatment, can be usefully added to dosimetric calculations, to compare different treatment schedules. The Linear-Quadratic model is the most used: radiobiological modelisation and Biologically Effective Dose (BED) is principal related dosimetric quantity. Therefore, the consideration of BED-volume histogram (BED-VHs) is a straightforward extension of DVHs. In practice, BED-VHs can help relative comparisons and optimisations in treatment planning when combined to dose-volume histograms. Since 1994 the dosimetric calculations for all the gynecological brachytherapy treatments are performed considering also DVHs and BED-VHs. In this presentation we show the methods of BEDVHs calculation, together with some typical results

  14. Automated intraoperative calibration for prostate cancer brachytherapy

    International Nuclear Information System (INIS)

    Kuiran Chen, Thomas; Heffter, Tamas; Lasso, Andras; Pinter, Csaba; Abolmaesumi, Purang; Burdette, E. Clif; Fichtinger, Gabor

    2011-01-01

    Purpose: Prostate cancer brachytherapy relies on an accurate spatial registration between the implant needles and the TRUS image, called ''calibration''. The authors propose a new device and a fast, automatic method to calibrate the brachytherapy system in the operating room, with instant error feedback. Methods: A device was CAD-designed and precision-engineered, which mechanically couples a calibration phantom with an exact replica of the standard brachytherapy template. From real-time TRUS images acquired from the calibration device and processed by the calibration system, the coordinate transformation between the brachytherapy template and the TRUS images was computed automatically. The system instantly generated a report of the target reconstruction accuracy based on the current calibration outcome. Results: Four types of validation tests were conducted. First, 50 independent, real-time calibration trials yielded an average of 0.57 ± 0.13 mm line reconstruction error (LRE) relative to ground truth. Second, the averaged LRE was 0.37 ± 0.25 mm relative to ground truth in tests with six different commercial TRUS scanners operating at similar imaging settings. Furthermore, testing with five different commercial stepper systems yielded an average of 0.29 ± 0.16 mm LRE relative to ground truth. Finally, the system achieved an average of 0.56 ± 0.27 mm target registration error (TRE) relative to ground truth in needle insertion tests through the template in a water tank. Conclusions: The proposed automatic, intraoperative calibration system for prostate cancer brachytherapy has achieved high accuracy, precision, and robustness.

  15. Orbital rhabdomyosarcoma of the child: the role of PDR brachytherapy in eye preservation

    International Nuclear Information System (INIS)

    Kovacs, G.; Rochels, R.; Mehdorn, H.M.; Werner, J.; Wilhelm, R.; Kohr, P.; Kimmig, B. N.

    1996-01-01

    Material and Methods: There were four children (8-7-5 years and(15(12)) months old) with recurrent/primary embryonal rhabdomyosarcoma treated with curative intention by peroperative PDR boost brachytherapy in combination with radio-chemotherapy and/or surgery. PDR brachytherapy according to the Kiel protocol: daily five pulses, two hours each, with 1 Gy on the reference isodose which is usually 2-3 mm close to the applicator surface. CT simulation based conformal treatment planning was carried out in each case. The implant was done intraoperatively using the free-hand plastic tube method, after a macroscopically complete excision of the tumor. Due to treatment planning individual target volume, eye with N, opticus and bone structures, as well as the applicators and other regions of interest were visualized. Manual volume optimisation was practiced and natural volumen-dose histograms were analysed in 'classic' graphic mode as well as in a special colour coded three-dimensional visualization in cine mode on the screen. One child received, three months before the recurrence was operated, 50 Gy hyperfractionated external beam radiation (2 Gy fractions) and was irradiated with 20 Gy brachytherapy in four days. The second patient received ten days after 20 Gy brachytherapy 32 Gy hyperfractionated external beam radiation. The third child (external beam treatment outside of our clinic), received conventional fractionated irradiation with 1.6 Gy fraction dose instead of a prescribed hyperfractionated external beam therapy and her brachytherapy dose was 25 Gy. At the (15(12)) months old child with primary embryonal rhabdomyosarcoma we applied 20 Gy brachytherapy and 24 Gy hyperfractionated external beam irradiation. All patients received multidrug chemotherapy according to the German Study Protocol (CWS-91). Results: Follow-up is 34, 28, 22, and 6 months for recurrent embryonal rhabdomyosarcoma patients (stand February 96). We observed at 9 months one rhabdomyosarcoma

  16. Clinical outcome and cosmetic results of conservative surgery plus radiation therapy in early stage breast cancer patients

    International Nuclear Information System (INIS)

    Jin Yening; Wang Yajie; Zhang Xiaoqing; Meng Yan; Li Rongqing; Shi Junyi

    2005-01-01

    Objective: To evaluate the clinical outcome and cosmetic results in early stage breast cancer patients treated with conservative surgery plus radiation therapy. Methods: From May 1995 to December 2002, 109 such patients were so treated. The post-operative radiotherapy consisted of whole-breast 6 MV linear accelerator irradiation with two tangential half-fields to a total dose of 45-52 Gy (mean 48.6 Gy), followed by a boost irradiation to the tumor bed. Among them, 79 patients received 10-12 Gy (DB) boost by interstitial implantation brachytherapy ( 192 Ir HDR, Nucletron), with single plane implantation for T1 and double plane implantation for T2 tumor. Thirty patients received 15 Gy boost by electron beam. Adjuvant/concurrent chemotherapy (CMF or CEF) and hormonotherapy were also used according to the patients' clinical characteristics. The cosmetic results were scored by both the doctor and the patients. Results: The overall actuarial 5-year survival was 93.8%, with local recurrence of 6.5%. No radiation-induced ulcer was observed in the breast except for acute inflammation at skin pinholes in 5 patients treated by interstitial implant brachytherapy. Among the 75 patients who had had breast examination, cosmetic result scored as good by patient and doctor were 81% and 87%, respectively. The good rate assessed by doctor in brachytherapy boost group and electron beam boost group were 81.2% (39/48) and 85.2% (23/27), There was no significant difference between these two boost techniques (P>0.05). Conclusions: Tumor bed boost irradiation by either brachytherapy or electron beam technique can provide satisfactory local control in early breast cancer treated with conservative surgery plus radiotherapy without increasing the side effects. There is no significant difference in cosmetic result between these two boost techniques. (authors)

  17. Biologically effective dose (BED) for interstitial seed implants containing a mixture of radionuclides with different half-lives

    International Nuclear Information System (INIS)

    Chen Zhe; Nath, Ravinder

    2003-01-01

    Purpose: To develop a tool for evaluating interstitial seed implants that contain a mixture of radionuclides with different half-lives and to demonstrate its utility by examining the clinical implications of prescribing to an isodose surface for such an implant. Methods and Materials: A linear-quadratic model for continuous low dose rate irradiation was developed for permanent implants containing a mixture of radionuclides. Using a generalized equation for the biologically effective dose (BED), the effects of cell proliferation and sublethal damage repair were examined systematically for implants containing a mixture of radionuclides. A head-and-neck permanent seed implant that contained a mixture of 125 I and 103 Pd seeds was used to demonstrate the utility of the generalized BED. Results: An equation of BED for implants containing a mixture of radionuclides with different half-lives was obtained. In such an implant, the effective cell kill was shown to depend strongly on the relative dose contributions from each radionuclide type; dose delivered by radionuclides with shorter half-life always resulted in more cell kill for any given sublethal damage repair and cell proliferation rates. Application of the BED formula to an implant containing a mixture of 125 I and 103 Pd seeds demonstrates that the conventional dose prescription to an isodose surface is not unique for such an implant. When the prescription dose was based on existing clinical experience of using 125 I seeds alone, mixing 103 Pd seeds with 125 I seeds would increase the cell kill. On the other hand, if the prescription dose were based on existing clinical experience of using 103 Pd seeds alone, mixing 125 I seeds with 103 Pd seeds in the same implant would create radiobiologically 'cold' spots (i.e., an increase in cell survival) at locations where a major portion of the prescription dose is contributed by the 125 I seeds. For fast-growing tumors, these 'cold' spots can become significant

  18. Radioablation of liver malignancies with interstitial high-dose-rate brachytherapy. Complications and risk factors

    Energy Technology Data Exchange (ETDEWEB)

    Mohnike, Konrad; Wolf, Steffen; Damm, Robert; Seidensticker, Max; Seidensticker, Ricarda; Fischbach, Frank; Pech, Maciej; Ricke, Jens [Otto-von-Guericke-Universitaet, Klinik fuer Radiologie und Nuklearmedizin, Universitaetsklinikum Magdeburg A.oe.R., Magdeburg (Germany); Peters, Nils; Hass, Peter; Gademann, Guenther [Otto-von-Guericke-Universitaet, Klinik fuer Strahlentherapie, Universitaetsklinikum Magdeburg A.oe.R., Magdeburg (Germany)

    2016-05-15

    To evaluate complications and identify risk factors for adverse events in patients undergoing high-dose-rate interstitial brachytherapy (iBT). Data from 192 patients treated in 343 CT- or MRI-guided interventions from 2006-2009 at our institution were analyzed. In 41 %, the largest tumor treated was ≥ 5 cm, 6 % of the patients had tumors ≥ 10 cm. Prior to iBT, 60 % of the patients had chemotherapy, 22 % liver resection, 19 % thermoablation or transarterial chemoembolization (TACE). Safety was the primary endpoint; survival data were obtained as the secondary endpoints. During follow-up, MRI or CT imaging was performed and clinical and laboratory parameters were obtained. The rate of major complications was below 5 %. Five major bleedings (1.5 %) occurred. The frequency of severe bleeding was significantly higher in patients with advanced liver cirrhosis. One patient developed signs of a nonclassic radiation-induced liver disease. In 3 patients, symptomatic gastrointestinal (GI) ulcers were detected. A dose exposure to the GI wall above 14 Gy/ml was a reliable threshold to predict ulcer formation. A combination of C-reactive protein ≥ 165 mg/l and/or leukocyte count ≥ 12.7 Gpt/l on the second day after the intervention predicted infection (sensitivity 90.0 %; specificity 92.8 %.) Two patients (0.6 %) died within 30 days. Median overall survival after the first liver treatment was 20.1 months for all patients and the local recurrence-free surviving proportion was 89 % after 12 months. Image-guided iBT yields a low rate of major complications and is effective. (orig.) [German] Evaluierung der Komplikationsrate und Identifizierung von Risikofaktoren fuer Komplikationen und Nebenwirkungen bei Patienten mit Lebermalignomen, die mit der hochdosierten interstitiellen Brachytherapie (iBT) behandelt wurden. Von 2006 bis 2009 wurden 192 Patienten in 343 CT- oder MRT-gefuehrten Interventionen behandelt und deren Daten ausgewertet. Der groesste behandelte Tumor war in

  19. Dose determination in breast tumor in brachytherapy using Iridium-192

    International Nuclear Information System (INIS)

    Okuno, S.F.

    1984-01-01

    Thermoluminescent dosimetry studies in vivo and in vitro aiming to determing radiation dose in the breast tumor, in brachytherapy using Iridium-192 was done. The correlation between radiation doses in tumor and external surface of the breast was investigated for correcting the time interval of radiation source implantation. (author) [pt

  20. Developing A Directional High-Dose Rate (d-HDR) Brachytherapy Source

    Science.gov (United States)

    Heredia, Athena Yvonne

    Conventional sources used in brachytherapy provide nearly isotropic or radially symmetric dose distributions. Optimizations of dose distributions have been limited to varied dwell times at specified locations within a given treatment volume, or manipulations in source position for seed implantation techniques. In years past, intensity modulated brachytherapy (IMBT) has been used to reduce the amount of radiation to surrounding sensitive structures in select intracavitary cases by adding space or partial shields. Previous work done by Lin et al., at the University of Wisconsin-Madison, has shown potential improvements in conformality for brachytherapy treatments using a directionally shielded low dose rate (LDR) source for treatments in breast and prostate. Directional brachytherapy sources irradiate approximately half of the radial angles around the source, and adequately shield a quarter of the radial angles on the opposite side, with sharp gradient zones between the treated half and shielded quarter. With internally shielded sources, the radiation can be preferentially emitted in such a way as to reduce toxicities in surrounding critical organs. The objective of this work is to present findings obtained in the development of a new directional high dose rate (d-HDR) source. To this goal, 103Pd (Z = 46) is reintroduced as a potential radionuclide for use in HDR brachytherapy. 103Pd has a low average photon energy (21 keV) and relatively short half -life (17 days), which is why it has historically been used in low dose rate applications and implantation techniques. Pd-103 has a carrier-free specific activity of 75000 Ci/g. Using cyclotron produced 103Pd, near carrier-free specific activities can be achieved, providing suitability for high dose rate applications. The evolution of the d-HDR source using Monte Carlo simulations is presented, along with dosimetric parameters used to fully characterize the source. In addition, a discussion on how to obtain elemental

  1. Reappraisal of the role of endobronchial brachytherapy in the management of lung cancer: 10 years' experience at the centre Antoine-Lacassagne

    International Nuclear Information System (INIS)

    Magne, N.; Benezery, K.; Marcie, S.; Lagrange, J.L.; Porsin, B.; Poudenx, M.; Otto, J.; Marcy, P.Y.; Benezery, K.; Lagrange, J.L.

    2003-01-01

    Intra-operative interstitial brachytherapy has been applied in the curative and palliative treatment of lung cancer. Implantation of radio-active sources offers an advantage over external irradiation because of the limited penetrability from source to prescription point, resulting in rapid dose fall-off and sparing of surrounding normal tissues. The aim of this study was to re-evaluate retrospectively the Antoine-Lacassagne cancer center experience in endobronchial brachytherapy by low dose rate (LDR) or high dose rate (HDR) and to design perspectives for the next decades. Evaluation was based on analysis of toxicities, response rates and survival. Materials and methods: From october 1989 to june 1999, 31 consecutive patients with bronchogenic carcinoma were treated. Thirteen and 18 patients received LDR and HDR, respectively. The mean age was 65 years (range 44 to 79 years). Inclusion criteria were, for palliative treatment, incurable endobronchial cancer, and for curative treatment, residual tumor in the margins after resection, or endobronchial tumor could not be treated surgically. Exclusion criteria were sites of lesion unsuitable for placement of the brachytherapy catheter. Evaluation of complications and clinical response were based on endoscopic evaluation one month after the last session and at less one year after the end of treatment. Eighty-seven courses have been performed: 65 by LDR and 22 by HDR. Thirty-six courses have been performed in the palliative group, 51 courses in the curative group. Seven patients among 31 presented acute complications and 18/31 late complications. Complete global response rate was 14/30 evaluable patients (47%). Mean overall Global survival was 23 months with a median follow-up of 3.5 years. These results confirm the efficacy of endobronchial brachytherapy as well as palliative or curative treatment, but the improvement of results will essentially depend on our capacity to better define our indications and underlie the

  2. A gEUD-based inverse planning technique for HDR prostate brachytherapy: Feasibility study

    Energy Technology Data Exchange (ETDEWEB)

    Giantsoudi, D. [Department of Radiological Sciences, University of Texas Health Sciences Center, San Antonio, Texas 78229 (United States); Department of Radiation Oncology, Francis H. Burr Proton Therapy Center, Boston, Massachusetts 02114 (United States); Baltas, D. [Department of Medical Physics and Engineering, Strahlenklinik, Klinikum Offenbach GmbH, 63069 Offenbach (Germany); Nuclear and Particle Physics Section, Physics Department, University of Athens, 15701 Athens (Greece); Karabis, A. [Pi-Medical Ltd., Athens 10676 (Greece); Mavroidis, P. [Department of Radiological Sciences, University of Texas Health Sciences Center, San Antonio, Texas 78299 and Department of Medical Radiation Physics, Karolinska Institutet and Stockholm University, 17176 (Sweden); Zamboglou, N.; Tselis, N. [Strahlenklinik, Klinikum Offenbach GmbH, 63069 Offenbach (Germany); Shi, C. [St. Vincent' s Medical Center, 2800 Main Street, Bridgeport, Connecticut 06606 (United States); Papanikolaou, N. [Department of Radiological Sciences, University of Texas Health Sciences Center, San Antonio, Texas 78299 (United States)

    2013-04-15

    Purpose: The purpose of this work was to study the feasibility of a new inverse planning technique based on the generalized equivalent uniform dose for image-guided high dose rate (HDR) prostate cancer brachytherapy in comparison to conventional dose-volume based optimization. Methods: The quality of 12 clinical HDR brachytherapy implants for prostate utilizing HIPO (Hybrid Inverse Planning Optimization) is compared with alternative plans, which were produced through inverse planning using the generalized equivalent uniform dose (gEUD). All the common dose-volume indices for the prostate and the organs at risk were considered together with radiobiological measures. The clinical effectiveness of the different dose distributions was investigated by comparing dose volume histogram and gEUD evaluators. Results: Our results demonstrate the feasibility of gEUD-based inverse planning in HDR brachytherapy implants for prostate. A statistically significant decrease in D{sub 10} or/and final gEUD values for the organs at risk (urethra, bladder, and rectum) was found while improving dose homogeneity or dose conformity of the target volume. Conclusions: Following the promising results of gEUD-based optimization in intensity modulated radiation therapy treatment optimization, as reported in the literature, the implementation of a similar model in HDR brachytherapy treatment plan optimization is suggested by this study. The potential of improved sparing of organs at risk was shown for various gEUD-based optimization parameter protocols, which indicates the ability of this method to adapt to the user's preferences.

  3. Polyethylene glycol hydrogel rectal spacer implantation in patients with prostate cancer undergoing combination high-dose-rate brachytherapy and external beam radiotherapy.

    Science.gov (United States)

    Yeh, Jekwon; Lehrich, Brandon; Tran, Carolyn; Mesa, Albert; Baghdassarian, Ruben; Yoshida, Jeffrey; Torrey, Robert; Gazzaniga, Michael; Weinberg, Alan; Chalfin, Stuart; Ravera, John; Tokita, Kenneth

    2016-01-01

    To present rectal toxicity rates in patients administered a polyethylene glycol (PEG) hydrogel rectal spacer in conjunction with combination high-dose-rate brachytherapy and external beam radiotherapy. Between February 2010 and April 2015, 326 prostate carcinoma patients underwent combination high-dose-rate brachytherapy of 16 Gy (average dose 15.5 Gy; standard deviation [SD] = 1.6 Gy) and external beam radiotherapy of 59.4 Gy (average dose 60.2 Gy; SD = 2.9 Gy). In conjunction with the radiation therapy regimen, each patient was injected with 10 mL of a PEG hydrogel in the anterior perirectal fat space. The injectable spacer (rectal spacer) creates a gap between the prostate and the rectum. The rectum is displaced from the radiation field, and rectal dose is substantially reduced. The goal is a reduction in rectal radiation toxicity. Clinical efficacy was determined by measuring acute and chronic rectal toxicity using the National Cancer Center Institute Common Terminology Criteria for Adverse Events v4.0 grading scheme. Median followup was 16 months. The mean anterior-posterior separation achieved was 1.6 cm (SD = 0.4 cm). Rates of acute Grade 1 and 2 rectal toxicity were 37.4% and 2.8%, respectively. There were no acute Grade 3/4 toxicities. Rates of late Grade 1, 2, and 3 rectal toxicity were 12.7%, 1.4%, and 0.7%, respectively. There were no late Grade 4 toxicities. PEG rectal spacer implantation is safe and well tolerated. Acute and chronic rectal toxicities are low despite aggressive dose escalation. Copyright © 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  4. WE-AB-BRA-12: Post-Implant Dosimetry in Prostate Brachytherapy by X-Ray and MRI Fusion

    Energy Technology Data Exchange (ETDEWEB)

    Park, S; Song, D; Lee, J [Johns Hopkins University, Baltimore, MD (United States); Le, Y [Indiana University, Indianapolis, IN (United States)

    2016-06-15

    Purpose: For post-implant dosimetric assessment after prostate brachytherapy, CT-MR fusion approach has been advocated due to the superior accuracy on both seeds localization and soft tissue delineation. However, CT deposits additional radiation to the patient, and seed identification in CT requires manual review and correction. In this study, we propose an accurate, low-dose, and cost-effective post-implant dosimetry approach based on X-ray and MRI. Methods: Implanted seeds are reconstructed using only three X-ray fluoroscopy images by solving a combinatorial optimization problem. The reconstructed seeds are then registered to MR images using an intensity-based points-to-volume registration. MR images are first pre-processed by geometric and Gaussian filtering, yielding smooth candidate seed-only images. To accommodate potential soft tissue deformation, our registration is performed in two steps, an initial affine followed by local deformable registrations. An evolutionary optimizer in conjunction with a points-to-volume similarity metric is used for the affine registration. Local prostate deformation and seed migration are then adjusted by the deformable registration step with external and internal force constraints. Results: We tested our algorithm on twenty patient data sets. For quantitative evaluation, we obtained ground truth seed positions by fusing the post-implant CT-MR images. Seeds were semi-automatically extracted from CT and manually corrected and then registered to the MR images. Target registration error (TRE) was computed by measuring the Euclidean distances from the ground truth to the closest registered X-ray seeds. The overall TREs (mean±standard deviation in mm) are 1.6±1.1 (affine) and 1.3±0.8 (affine+deformable). The overall computation takes less than 1 minute. Conclusion: It has been reported that the CT-based seed localization error is ∼1.6mm and the seed localization uncertainty of 2mm results in less than 5% deviation of prostate D

  5. WE-AB-BRA-12: Post-Implant Dosimetry in Prostate Brachytherapy by X-Ray and MRI Fusion

    International Nuclear Information System (INIS)

    Park, S; Song, D; Lee, J; Le, Y

    2016-01-01

    Purpose: For post-implant dosimetric assessment after prostate brachytherapy, CT-MR fusion approach has been advocated due to the superior accuracy on both seeds localization and soft tissue delineation. However, CT deposits additional radiation to the patient, and seed identification in CT requires manual review and correction. In this study, we propose an accurate, low-dose, and cost-effective post-implant dosimetry approach based on X-ray and MRI. Methods: Implanted seeds are reconstructed using only three X-ray fluoroscopy images by solving a combinatorial optimization problem. The reconstructed seeds are then registered to MR images using an intensity-based points-to-volume registration. MR images are first pre-processed by geometric and Gaussian filtering, yielding smooth candidate seed-only images. To accommodate potential soft tissue deformation, our registration is performed in two steps, an initial affine followed by local deformable registrations. An evolutionary optimizer in conjunction with a points-to-volume similarity metric is used for the affine registration. Local prostate deformation and seed migration are then adjusted by the deformable registration step with external and internal force constraints. Results: We tested our algorithm on twenty patient data sets. For quantitative evaluation, we obtained ground truth seed positions by fusing the post-implant CT-MR images. Seeds were semi-automatically extracted from CT and manually corrected and then registered to the MR images. Target registration error (TRE) was computed by measuring the Euclidean distances from the ground truth to the closest registered X-ray seeds. The overall TREs (mean±standard deviation in mm) are 1.6±1.1 (affine) and 1.3±0.8 (affine+deformable). The overall computation takes less than 1 minute. Conclusion: It has been reported that the CT-based seed localization error is ∼1.6mm and the seed localization uncertainty of 2mm results in less than 5% deviation of prostate D

  6. Monte Carlo study of LDR seed dosimetry with an application in a clinical brachytherapy breast implant.

    Science.gov (United States)

    Furstoss, C; Reniers, B; Bertrand, M J; Poon, E; Carrier, J-F; Keller, B M; Pignol, J P; Beaulieu, L; Verhaegen, F

    2009-05-01

    A Monte Carlo (MC) study was carried out to evaluate the effects of the interseed attenuation and the tissue composition for two models of 125I low dose rate (LDR) brachytherapy seeds (Medi-Physics 6711, IBt InterSource) in a permanent breast implant. The effect of the tissue composition was investigated because the breast localization presents heterogeneities such as glandular and adipose tissue surrounded by air, lungs, and ribs. The absolute MC dose calculations were benchmarked by comparison to the absolute dose obtained from experimental results. Before modeling a clinical case of an implant in heterogeneous breast, the effects of the tissue composition and the interseed attenuation were studied in homogeneous phantoms. To investigate the tissue composition effect, the dose along the transverse axis of the two seed models were calculated and compared in different materials. For each seed model, three seeds sharing the same transverse axis were simulated to evaluate the interseed effect in water as a function of the distance from the seed. A clinical study of a permanent breast 125I implant for a single patient was carried out using four dose calculation techniques: (1) A TG-43 based calculation, (2) a full MC simulation with realistic tissues and seed models, (3) a MC simulation in water and modeled seeds, and (4) a MC simulation without modeling the seed geometry but with realistic tissues. In the latter, a phase space file corresponding to the particles emitted from the external surface of the seed is used at each seed location. The results were compared by calculating the relevant clinical metrics V85, V100, and V200 for this kind of treatment in the target. D90 and D50 were also determined to evaluate the differences in dose and compare the results to the studies published for permanent prostate seed implants in literature. The experimental results are in agreement with the MC absolute doses (within 5% for EBT Gafchromic film and within 7% for TLD-100

  7. Computed Tomography–Guided Interstitial High-Dose-Rate Brachytherapy in Combination With Regional Positive Lymph Node Intensity-Modulated Radiation Therapy in Locally Advanced Peripheral Non–Small Cell Lung Cancer: A Phase 1 Clinical Trial

    Energy Technology Data Exchange (ETDEWEB)

    Xiang, Li; Zhang, Jian-wen; Lin, Sheng; Luo, Hui-Qun; Wen, Qing-Lian; He, Li-Jia; Shang, Chang-Ling; Ren, Pei-Rong; Yang, Hong-Ru; Pang, Hao-Wen; Yang, Bo; He, Huai-Lin [Department of Oncology, Affiliated Hospital of Luzhou Medical College, Luzhou (China); Chen, Yue, E-mail: chenyue5523@126.com [Department of Nuclear Medicine, Affiliated Hospital of Luzhou Medical College, Luzhou (China); Wu, Jing-Bo, E-mail: wjb6147@163.com [Department of Oncology, Affiliated Hospital of Luzhou Medical College, Luzhou (China)

    2015-08-01

    Purpose: To assess the technical safety, adverse events, and efficacy of computed tomography (CT)-guided interstitial high-dose-rate (HDR) brachytherapy in combination with regional positive lymph node intensity modulated radiation therapy in patients with locally advanced peripheral non–small cell lung cancer (NSCLC). Methods and Materials: Twenty-six patients with histologically confirmed NSCLC were enrolled in a prospective, officially approved phase 1 trial. Primary tumors were treated with HDR brachytherapy. A single 30-Gy dose was delivered to the 90% isodose line of the gross lung tumor volume. A total dose of at least 70 Gy was administered to the 95% isodose line of the planning target volume of malignant lymph nodes using 6-MV X-rays. The patients received concurrent or sequential chemotherapy. We assessed treatment efficacy, adverse events, and radiation toxicity. Results: The median follow-up time was 28 months (range, 7-44 months). There were 3 cases of mild pneumothorax but no cases of hemothorax, dyspnea, or pyothorax after the procedure. Grade 3 or 4 acute hematologic toxicity was observed in 5 patients. During follow-up, mild fibrosis around the puncture point was observed on the CT scans of 2 patients, but both patients were asymptomatic. The overall response rates (complete and partial) for the primary mass and positive lymph nodes were 100% and 92.3%, respectively. The 1-year and 2-year overall survival (OS) rates were 90.9% and 67%, respectively, with a median OS of 22.5 months. Conclusion: Our findings suggest that HDR brachytherapy is safe and feasible for peripheral locally advanced NSCLC, justifying a phase 2 clinical trial.

  8. Brachytherapy optimization using radiobiological-based planning for high dose rate and permanent implants for prostate cancer treatment

    Science.gov (United States)

    Seeley, Kaelyn; Cunha, J. Adam; Hong, Tae Min

    2017-01-01

    We discuss an improvement in brachytherapy--a prostate cancer treatment method that directly places radioactive seeds inside target cancerous regions--by optimizing the current standard for delivering dose. Currently, the seeds' spatiotemporal placement is determined by optimizing the dose based on a set of physical, user-defined constraints. One particular approach is the ``inverse planning'' algorithms that allow for tightly fit isodose lines around the target volumes in order to reduce dose to the patient's organs at risk. However, these dose distributions are typically computed assuming the same biological response to radiation for different types of tissues. In our work, we consider radiobiological parameters to account for the differences in the individual sensitivities and responses to radiation for tissues surrounding the target. Among the benefits are a more accurate toxicity rate and more coverage to target regions for planning high-dose-rate treatments as well as permanent implants.

  9. Physics and quality assurance for brachytherapy - Part I: High dose rates

    International Nuclear Information System (INIS)

    Anderson, Lowell L.

    1997-01-01

    Purpose: To review the physical aspects of high dose rate (HDR) brachytherapy, including commissioning and quality assurance, source calibration and dose distribution measurements, and treatment planning methods. Following the introduction of afterloading in brachytherapy, development efforts to make it 'remote' culminated in 1964 with the near-simultaneous appearance of remote afterloaders in five major medical centers. Four of these machines were 'high dose rate', three employing 60Co and one (the GammaMed) using a single, cable-mounted 192Ir source. Stepping-motor source control was added to the GammaMed in 1974, making it the precursor of modern remote afterloaders, which are now suitable for interstitial, well as intracavitary brachytherapy by virtue of small source-diameter and indexer-accessed multiple channels. Because the 192Ir sources currently used in HDR remote afterloaders are supplied at a nominal air-kerma strength of 11.4 cGy cm2 s-1 (10 Ci), are not collimated in clinical use, and emit a significant fraction (15%) of photons at energies greater than 600 keV, shielding and facility design must be undertaken as carefully and thoroughly as for external beam installations. Licensing requirements of regulatory agencies must be met with respect both to maximum permissible dose limits and to the existence and functionality of safety devices (door interlocks, radiation monitors, etc.). Commissioning and quality assurance procedures that must be documented for HDR remote afterloading relate to (1) machine, applicator, guide-tube, and facility functionality checks, (2) source calibration, (3) emergency response readiness, (4) planning software evaluation, and (5) independent checks of clinical dose calculations. Source calibration checks must be performed locally, either by in-air measurement of air kerma strength or with a well ionization chamber calibrated (by an accredited standards laboratory) against an in-air measurement of air kerma strength for the

  10. Experimental radiotherapy of the R1H rhabdomyosarcoma of the rat: combined use of interstitial iodine-125-brachytherapy and fractionated X-irradiation

    International Nuclear Information System (INIS)

    Doll, D.

    1995-01-01

    The study described here investigated into the therapeutic effects that split-dose x-radiation combined with interstitial iodine-125 brachytherapy would have on two different lines of the R1H rhabdomyosarcoma of the rat. The following parameters were examined: local tumour control rate; growth delay; net growth delay; position, movement and loss of seeds; tumour shape. The following results were obtained: The local tumour control rate for tumours externally treated with two seeds was by 42 Gy higher than that determined for the group treated with external irradiation alone. A procedure was developed to calculate the most appropriate distance for the seeds on the basis of tumour axes and volumes. The relationship between growth delay and mean maximum distance of the seed from the tumour margin could be ascertained on a quantitative basis. The influence of the tumour shape on the result of treatment was confirmed. Although the seeds were still active at the time of recidivation and treatment was not yet terminated, it was possible to show that the tumour bed effect, which tends to distort the growth delay calculations and may even occur in externally treated seed animals, could largely be avoided in the evaluation of this study. (orig./MG) [de

  11. Radiation exposure after permanent prostate brachytherapy

    International Nuclear Information System (INIS)

    Cattani, Federica; Vavassori, Andrea; Polo, Alfredo; Rondi, Elena; Cambria, Raffaella; Orecchia, Roberto; Tosi, Giampiero

    2006-01-01

    Background and purpose: Limited information is available on the true radiation exposure and associated risks for the relatives of the patients submitted to prostate brachytherapy with permanent implant of radioactive sources and for any other people coming into contact with them. In order to provide appropriate information, we analyzed the radiation exposure data from 216 prostate cancer patients who underwent 125 I or 103 Pd implants at the European Institute of Oncology of Milan, Italy. Patients and methods: Between October 1999 and October 2004, 216 patients with low risk prostate carcinoma were treated with 125 I (200 patients) or 103 Pd (16 patients) permanent seed implantation. One day after the procedure, radiation exposure measurements around the patients were performed using an ionization chamber survey meter (Victoreen RPO-50) calibrated in dose rate at an accredited calibration center (calibration Centre SIT 104). Results: The mean dose rate at the posterior skin surface (gluteal region) following 125 I implants was 41.3 μSv/h (range: 6.2-99.4 μSv/h) and following 103 Pd implants was 18.9 μSv/h (range 5.0-37.3 μSv/h). The dose rate at 50 cm from the skin decreased to the mean value of 6.4 μSv/h for the 125 I implants and to the mean value of 1.7 μSv/h for the 103 Pd implants. Total times required to reach the annual dose limit (1 mSv/year) recommended for the general population by the European Directive 96/29/Euratom and by the Italian law (Decreto Legislativo 241/2000) at a distance of 50 cm from the posterior skin surface of the implanted patient would be 7.7 and 21.6 days for 125 I and for 103 Pd. Good correlation between the measured dose rates and both the total implanted activity and the distance between the most posteriorly implanted seed and the skin surface of the patients was found. Conclusions: Our data show that the dose rates at 50 cm away from the prostate brachytherapy patients are very low and that the doses possibly absorbed by the

  12. Establishing High-Quality Prostate Brachytherapy Using a Phantom Simulator Training Program

    Energy Technology Data Exchange (ETDEWEB)

    Thaker, Nikhil G. [Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States); Kudchadker, Rajat J. [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States); Swanson, David A. [Department of Urology, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States); Albert, Jeffrey M. [Department of Radiation Oncology, Banner Health, Loveland/Greeley, Colorado (United States); Mahmood, Usama; Pugh, Thomas J.; Boehling, Nicholas S. [Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States); Bruno, Teresa L. [Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States); Prestidge, Bradley R. [Department of Radiation Oncology, Bon Secours Health System, Norfolk, Virginia (United States); Crook, Juanita M. [Department of Radiation Oncology, Cancer Center for the Southern Interior, Kelowna, British Columbia (Canada); Cox, Brett W.; Potters, Louis [Department of Radiation Medicine, North Shore-LIJ Health System, New Hyde Park, New York (United States); Moran, Brian J. [Chicago Prostate Center, Westmont, Illinois (United States); Keyes, Mira [Department of Radiation Oncology, British Columbia Cancer Agency, Vancouver Center, Vancouver, British Columbia (Canada); Kuban, Deborah A. [Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States); Frank, Steven J., E-mail: sjfrank@mdanderson.org [Department of Radiation Oncology, The University of Texas MD Anderson Cancer Center, Houston, Texas (United States)

    2014-11-01

    Purpose: To design and implement a unique training program that uses a phantom-based simulator to teach the process of prostate brachytherapy (PB) quality assurance and improve the quality of education. Methods and Materials: Trainees in our simulator program were practicing radiation oncologists, radiation oncology residents, and fellows of the American Brachytherapy Society. The program emphasized 6 core areas of quality assurance: patient selection, simulation, treatment planning, implant technique, treatment evaluation, and outcome assessment. Using the Iodine 125 ({sup 125}I) preoperative treatment planning technique, trainees implanted their ultrasound phantoms with dummy seeds (ie, seeds with no activity). Pre- and postimplant dosimetric parameters were compared and correlated using regression analysis. Results: Thirty-one trainees successfully completed the simulator program during the period under study. The mean phantom prostate size, number of seeds used, and total activity were generally consistent between trainees. All trainees met the V100 >95% objective both before and after implantation. Regardless of the initial volume of the prostate phantom, trainees' ability to cover the target volume with at least 100% of the dose (V100) was not compromised (R=0.99 pre- and postimplant). However, the V150 had lower concordance (R=0.37) and may better reflect heterogeneity control of the implant process. Conclusions: Analysis of implants from this phantom-based simulator shows a high degree of consistency between trainees and uniformly high-quality implants with respect to parameters used in clinical practice. This training program provides a valuable educational opportunity that improves the quality of PB training and likely accelerates the learning curve inherent in PB. Prostate phantom implantation can be a valuable first step in the acquisition of the required skills to safely perform PB.

  13. Establishing High-Quality Prostate Brachytherapy Using a Phantom Simulator Training Program

    International Nuclear Information System (INIS)

    Thaker, Nikhil G.; Kudchadker, Rajat J.; Swanson, David A.; Albert, Jeffrey M.; Mahmood, Usama; Pugh, Thomas J.; Boehling, Nicholas S.; Bruno, Teresa L.; Prestidge, Bradley R.; Crook, Juanita M.; Cox, Brett W.; Potters, Louis; Moran, Brian J.; Keyes, Mira; Kuban, Deborah A.; Frank, Steven J.

    2014-01-01

    Purpose: To design and implement a unique training program that uses a phantom-based simulator to teach the process of prostate brachytherapy (PB) quality assurance and improve the quality of education. Methods and Materials: Trainees in our simulator program were practicing radiation oncologists, radiation oncology residents, and fellows of the American Brachytherapy Society. The program emphasized 6 core areas of quality assurance: patient selection, simulation, treatment planning, implant technique, treatment evaluation, and outcome assessment. Using the Iodine 125 ( 125 I) preoperative treatment planning technique, trainees implanted their ultrasound phantoms with dummy seeds (ie, seeds with no activity). Pre- and postimplant dosimetric parameters were compared and correlated using regression analysis. Results: Thirty-one trainees successfully completed the simulator program during the period under study. The mean phantom prostate size, number of seeds used, and total activity were generally consistent between trainees. All trainees met the V100 >95% objective both before and after implantation. Regardless of the initial volume of the prostate phantom, trainees' ability to cover the target volume with at least 100% of the dose (V100) was not compromised (R=0.99 pre- and postimplant). However, the V150 had lower concordance (R=0.37) and may better reflect heterogeneity control of the implant process. Conclusions: Analysis of implants from this phantom-based simulator shows a high degree of consistency between trainees and uniformly high-quality implants with respect to parameters used in clinical practice. This training program provides a valuable educational opportunity that improves the quality of PB training and likely accelerates the learning curve inherent in PB. Prostate phantom implantation can be a valuable first step in the acquisition of the required skills to safely perform PB

  14. CT-image based conformal brachytherapy of breast cancer. The significance of semi-3-D and 3-D treatment planning

    International Nuclear Information System (INIS)

    Polgar, C.; Major, T.; Somogyi, A.; Takacsi-Nagy, Z.; Mangel, L.C.; Fodor, J.; Nemeth, G.; Forrai, G.; Sulyok, Z.

    2000-01-01

    In 103 patients with T1-2, N0-1 breast cancer the tumor bed was clipped during breast conserving surgery. Fifty-two of them received boost brachytherapy after 46 to 50 Gy teletherapy and 51 patients were treated with brachytherapy alone via flexible implant tubes. Single double and triple plane implant was used in 6,89 and 8 cases, respectively. The dose of boost brachytherapy and sole brachytherapy prescribed to dose reference points was 3 times 4.75 Gy and 7 times 5.2 Gy, respectively. The positions of dose reference points varied according to the level (2-D, semi-3-D and 3-D) of treatment planning performed. The treatment planning was based on the 3-D reconstruction of the surgical clips, implant tubes and skin points. In all cases the implantations were planned with a semi-3-D technique aided by simulator. In 10 cases a recently developed CT-guided 3-D planning system was used. The semi-3D and 3-D treatment plans were compared to hypothetical 2-D plans using dose-volume histograms and dose non-uniformity ratios. The values of mean central dose, mean skin dose, minimal clip dose, proportion of underdosaged clips and mean target surface dose were evaluated. The accuracy of tumor bed localization and the conformity of planning target volume and treated volume were also analyzed in each technique. Results: With the help of conformal semi-3D and 3D brachytherapy planning we could define reference dose points, active source positions and dwell times individually. This technique decreased the mean skin dose with 22.2% and reduced the possibility of geographical miss. We could achieve the best conformity between the planning target volume and the treated volume with the CT-image based 3-D treatment planning, at the cost of worse dose homogeneity. The mean treated volume was reduced by 25.1% with semi-3-D planning, however, its was increased by 16.2% with 3-D planning, compared to the 2-D planning. (orig.) [de

  15. Neodadjuvante und adjuvante Kurzzeit-Hormontherapie in Kombination mit konformaler HDR-Brachytherapie beim Prostatakarzinom

    Directory of Open Access Journals (Sweden)

    Martin T

    2004-01-01

    Full Text Available Zielsetzung: Auswertung der Behandlungsergebnisse der neoadjuvanten und adjuvanten Kurzzeit-Hormontherapie kombiniert mit konformaler HDR-Brachytherapie und externer Radiotherapie beim Prostatakarzinom. Patienten und Methoden: Von 01/97 bis 09/99 behandelten wir 102 Patienten mit Prostatakarzinomen im Stadium T1–3 N0 M0. Im Stadium T1–2 befanden sich 71, im Stadium T3 31 Patienten. Der mediane prätherapeutische PSA-Wert betrug 15,3 ng/ml. Nach ultraschallgesteuerter transrektaler Implantation von vier Afterloadingnadeln erfolgte die CT-gestützte 3D-Brachytherapie- Planung. Alle Patienten erhielten vier HDR-Implantate mit einer Referenzdosis von 5 Gy oder 7 Gy pro Implantat. Die Zeit zwischen jedem Implantat betrug jeweils 14 Tage. Nach der Brachytherapie folgte die externe Radiotherapie bis 39,6 Gy oder 45,0 Gy. Alle Patienten erhielten eine neoadjuvante und adjuvante Kurzzeit-Hormontherapie, die 2–19 Monate vor der Brachytherapie eingeleitet und 3 Monate nach Abschluß der externen Radiotherapie abgesetzt wurde (mediane Dauer: 9 Monate. Ergebnisse: Die mediane Nachbeobachtungszeit war 2,6 Jahre (range: 2,0–4,1 Jahre. Die biochemische Kontrollrate betrug 82 % nach 3 Jahren. Bei 14/102 Patienten registrierten wir ein biochemisches Rezidiv, bei 5/102 Patienten ein klinisches Rezidiv. Das Gesamtüberleben betrug 90 %, das krankheitsspezifische Überleben 98,0 % nach 3 Jahren. Ein Patient entwickelte eine prostato-urethro-rektale Fistel als späte Grad 4-Toxizität. Akute Grad-3 Toxizitäten traten bei 4 %, späte Grad-3 Toxizitäten bei 5 % der Patienten auf. Schlußfolgerung: Die neoadjuvante und adjuvante Kurzzeit-Hormontherapie kombiniert mit konformaler HDR-Brachytherapie und externer Radiotherapie erweist sich als sichere und wirksame Behandlungsmodalität beim Prostatakarzinom mit minimalen behandlungsbedingten Toxizitäten und einer vielversprechenden biochemischen Kontrollrate nach medianer Nachbeobachtungszeit von 2,6 Jahren.

  16. Aspects of an automatic system of implants of radioactive seeds and anatomic object simulator for tests in prostate brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Silva, Leonardo S.M.; Braga, Viviane V.B.; Campos, Tarcísio P. R. de, E-mail: leonardosantiago.lsms@gmail.com, E-mail: vitoriabraga06@gmail.com, E-mail: tprcampos@yahoo.com.br [Universidade Federal de Minas Gerais (PCTN/UFMG), Belo Horizonte (Brazil). Pós-Graduação em Ciências e Técnicas Nucleares. Departamento de Engenharia Nuclear

    2017-07-01

    This work presents the state of the art of the research and development of an automatic radioactive seed implantation system (PSIS). PSIS may assist in the procedure of testing permanent implants in the prostate. These tests will be important in measurements of absorbed doses in the pelvic structures, involving the organs and tissues at risk to improve planning, seed positioning and dosimetry. The automated Prostate Seed Implant System (PSIS) has been designed to meet operational needs, which offers the freedom of positioning of the brachytherapy needle within the treatment area and ensures repeatability and fidelity to the planned treatment. Both the ultrasound probe and the seed implant needle are driven by step motors, Atmega microcontroller, bearings, aluminum shafts and a GUI (Graphical User Interface). Movement of both the probe and the needle holder was performed by fixed spindle on a threaded rod rushing to the step motors by a coupling. The step motors used to move the system consist of step motors used in CNC (Computer Numeric Control) machine. The choice of these engines occurred due to the precision in the movements that can be obtained with these types of motors. The ultrasound probe serves to help, through the images acquired during the longitudinal movement, to monitor the application of the seeds. The parts that make up the system infrastructure were made of aluminum and translucent acrylic and cylindrical aluminum bars of different diameters. All these pieces were fixed and adjusted trough screws, washers, nuts and adhesive to metal, composing the final prototype of the PSIS. The project was developed and the PSIS prototype was assembled. The prototype presented acceptable operating characteristics for prostate implants. The advantage of this system is the automation of the application that provides an accurate positioning and movement of both probe and seed application. In addition to this study, seeds implantation tests will be performed, and

  17. Aspects of an automatic system of implants of radioactive seeds and anatomic object simulator for tests in prostate brachytherapy

    International Nuclear Information System (INIS)

    Silva, Leonardo S.M.; Braga, Viviane V.B.; Campos, Tarcísio P. R. de

    2017-01-01

    This work presents the state of the art of the research and development of an automatic radioactive seed implantation system (PSIS). PSIS may assist in the procedure of testing permanent implants in the prostate. These tests will be important in measurements of absorbed doses in the pelvic structures, involving the organs and tissues at risk to improve planning, seed positioning and dosimetry. The automated Prostate Seed Implant System (PSIS) has been designed to meet operational needs, which offers the freedom of positioning of the brachytherapy needle within the treatment area and ensures repeatability and fidelity to the planned treatment. Both the ultrasound probe and the seed implant needle are driven by step motors, Atmega microcontroller, bearings, aluminum shafts and a GUI (Graphical User Interface). Movement of both the probe and the needle holder was performed by fixed spindle on a threaded rod rushing to the step motors by a coupling. The step motors used to move the system consist of step motors used in CNC (Computer Numeric Control) machine. The choice of these engines occurred due to the precision in the movements that can be obtained with these types of motors. The ultrasound probe serves to help, through the images acquired during the longitudinal movement, to monitor the application of the seeds. The parts that make up the system infrastructure were made of aluminum and translucent acrylic and cylindrical aluminum bars of different diameters. All these pieces were fixed and adjusted trough screws, washers, nuts and adhesive to metal, composing the final prototype of the PSIS. The project was developed and the PSIS prototype was assembled. The prototype presented acceptable operating characteristics for prostate implants. The advantage of this system is the automation of the application that provides an accurate positioning and movement of both probe and seed application. In addition to this study, seeds implantation tests will be performed, and

  18. Prostate brachytherapy seed migration to the heart seen on cardiovascular computed tomographic angiography

    Directory of Open Access Journals (Sweden)

    Shilpa Sachdeva, MD

    2017-03-01

    Full Text Available Brachytherapy consists of placing radioactive sources into or adjacent to tumors, to deliver conformal radiation treatment. The technique is used for treatment of primary malignancies and for salvage in recurrent disease. Permanent prostate brachytherapy seeds are small metal implants containing radioactive sources of I-125, Pd-103, or Cs-131 encased in a titanium shell. They can embolize through the venous system to the lungs or heart and subsequently be detected by cardiovascular computed tomography. Cardiovascular imagers should be aware of the appearance of migrated seeds, as their presence in the chest is generally benign, so that unnecessary worry and testing are avoided. We report a case of a patient who underwent brachytherapy for prostate cancer and developed a therapeutic seeds embolus to the right ventricle.

  19. CT-image based conformal brachytherapy of breast cancer. The significance of semi-3-D and 3-D treatment planning

    Energy Technology Data Exchange (ETDEWEB)

    Polgar, C.; Major, T.; Somogyi, A.; Takacsi-Nagy, Z.; Mangel, L.C.; Fodor, J.; Nemeth, G. [Orszagos Onkologiai Intezet, Budapest (Hungary). Dept. of Radiotherapy; Forrai, G. [Haynal Imre Univ. of Health Sciences, Budapest (Hungary). Dept. of Radiology; Sulyok, Z. [Orszagos Onkologiai Intezet, Budapest (Hungary). Dept. of Surgery

    2000-03-01

    In 103 patients with T1-2, N0-1 breast cancer the tumor bed was clipped during breast conserving surgery. Fifty-two of them received boost brachytherapy after 46 to 50 Gy teletherapy and 51 patients were treated with brachytherapy alone via flexible implant tubes. Single double and triple plane implant was used in 6,89 and 8 cases, respectively. The dose of boost brachytherapy and sole brachytherapy prescribed to dose reference points was 3 times 4.75 Gy and 7 times 5.2 Gy, respectively. The positions of dose reference points varied according to the level (2-D, semi-3-D and 3-D) of treatment planning performed. The treatment planning was based on the 3-D reconstruction of the surgical clips, implant tubes and skin points. In all cases the implantations were planned with a semi-3-D technique aided by simulator. In 10 cases a recently developed CT-guided 3-D planning system was used. The semi-3D and 3-D treatment plans were compared to hypothetical 2-D plans using dose-volume histograms and dose non-uniformity ratios. The values of mean central dose, mean skin dose, minimal clip dose, proportion of underdosaged clips and mean target surface dose were evaluated. The accuracy of tumor bed localization and the conformity of planning target volume and treated volume were also analyzed in each technique. Results: With the help of conformal semi-3D and 3D brachytherapy planning we could define reference dose points, active source positions and dwell times individually. This technique decreased the mean skin dose with 22.2% and reduced the possibility of geographical miss. We could achieve the best conformity between the planning target volume and the treated volume with the CT-image based 3-D treatment planning, at the cost of worse dose homogeneity. The mean treated volume was reduced by 25.1% with semi-3-D planning, however, its was increased by 16.2% with 3-D planning, compared to the 2-D planning. (orig.) [German] Bei 103 Patientinnen mit Mammakarzinom der Stadien T1

  20. Rescue treatment with interstitial brachytherapy irradiation re very low dose rate iridium-192 (UBT) in inoperable tumors of the oral cavity, oropharynx and nodal: experience of 28 cases in the Gustave-Roussy Institute in Paris

    International Nuclear Information System (INIS)

    Quarneti, A.; Cordova, A.; Barrios, E.; Bonomi, M.; Haie-Meder, C.; Gerbaulet, A.; Eschwege, F.

    2004-01-01

    Purpose: A retrospective analysis of the evolution of 28 patients was performed local recurrences, second tumors and advanced disease in neck nodes in territory previously irradiated, which were re-irradiated using interstitial brachytherapy Ir-192 at very low dose rate (UBT) in the Gustave-Roussy Institute in Paris. Material and Methods: A series of 28 who had received radiation therapy is reported as part of heir initial treatment. 17 patients were treated for local recurrences or second tumors while 11 patients had presented nodal disease. All of them were inoperable. So were treated with interstitial brachytherapy with Ir-192 wires at very low rate dose (UBT), plastic tube technique, re-irradiation regime between 1978 and 1988 Gustave Roussy Institute. Two groups were considered. Group 1 included 17 patients with local recurrences, lesion progression and second tumors. Group 2 included 11 patients with metastatic nodal disease. The mean treatment volume was 45.25 cc, the average dose was 65 Gy, and the average treatment time between the first treatment and re irradiation was 56 months. The average duration of treatment was 14.6 days with a average dose rate of 0.18 Gy / h. After loading technique was used in plastic tubes. They were previously performed to load the simulation with orthogonal plates, false sources and provisional dosimetry. Late toxicity was assessed according to the RTOG score. Local control rates were studied complications and survive on some factors of possible prognostic significance. The statistical analysis of significance was performed by the method and log rank test were prepared survival curves and disease-free survival by Kaplan-Meier. Results: 2 groups were analyzed separately. In group 1, procedures were performed 17 UBTD and method of low dose rate (LDR). 10 of 17 patients achieved complete responses. The patient that the procedure was performed at low dose rate also achieved a complete response. In 3 cases, no response is not

  1. SU-E-T-123: Anomalous Altitude Effect in Permanent Implant Brachytherapy Seeds

    International Nuclear Information System (INIS)

    Watt, E; Spencer, DP; Meyer, T

    2015-01-01

    Purpose: Permanent seed implant brachytherapy procedures require the measurement of the air kerma strength of seeds prior to implant. This is typically accomplished using a well-type ionization chamber. Previous measurements (Griffin et al., 2005; Bohm et al., 2005) of several low-energy seeds using the air-communicating HDR 1000 Plus chamber have demonstrated that the standard temperature-pressure correction factor, P TP , may overcompensate for air density changes induced by altitude variations by up to 18%. The purpose of this work is to present empirical correction factors for two clinically-used seeds (IsoAid ADVANTAGE™ 103 Pd and Nucletron selectSeed 125 I) for which empirical altitude correction factors do not yet exist in the literature when measured with the HDR 1000 Plus chamber. Methods: An in-house constructed pressure vessel containing the HDR 1000 Plus well chamber and a digital barometer/thermometer was pumped or evacuated, as appropriate, to a variety of pressures from 725 to 1075 mbar. Current measurements, corrected with P TP , were acquired for each seed at these pressures and normalized to the reading at ‘standard’ pressure (1013.25 mbar). Results: Measurements in this study have shown that utilization of P TP can overcompensate in the corrected current reading by up to 20% and 17% for the IsoAid Pd-103 and the Nucletron I-125 seed respectively. Compared to literature correction factors for other seed models, the correction factors in this study diverge by up to 2.6% and 3.0% for iodine (with silver) and palladium respectively, indicating the need for seed-specific factors. Conclusion: The use of seed specific altitude correction factors can reduce uncertainty in the determination of air kerma strength. The empirical correction factors determined in this work can be applied in clinical quality assurance measurements of air kerma strength for two previously unpublished seed designs (IsoAid ADVANTAGE™ 103 Pd and Nucletron selectSeed 125 I

  2. SU-E-T-123: Anomalous Altitude Effect in Permanent Implant Brachytherapy Seeds

    Energy Technology Data Exchange (ETDEWEB)

    Watt, E; Spencer, DP; Meyer, T [University of Calgary and Tom Baker Cancer Centre, Calgary, AB (Canada)

    2015-06-15

    Purpose: Permanent seed implant brachytherapy procedures require the measurement of the air kerma strength of seeds prior to implant. This is typically accomplished using a well-type ionization chamber. Previous measurements (Griffin et al., 2005; Bohm et al., 2005) of several low-energy seeds using the air-communicating HDR 1000 Plus chamber have demonstrated that the standard temperature-pressure correction factor, P{sub TP}, may overcompensate for air density changes induced by altitude variations by up to 18%. The purpose of this work is to present empirical correction factors for two clinically-used seeds (IsoAid ADVANTAGE™ {sup 103}Pd and Nucletron selectSeed {sup 125}I) for which empirical altitude correction factors do not yet exist in the literature when measured with the HDR 1000 Plus chamber. Methods: An in-house constructed pressure vessel containing the HDR 1000 Plus well chamber and a digital barometer/thermometer was pumped or evacuated, as appropriate, to a variety of pressures from 725 to 1075 mbar. Current measurements, corrected with P{sub TP}, were acquired for each seed at these pressures and normalized to the reading at ‘standard’ pressure (1013.25 mbar). Results: Measurements in this study have shown that utilization of P{sub TP} can overcompensate in the corrected current reading by up to 20% and 17% for the IsoAid Pd-103 and the Nucletron I-125 seed respectively. Compared to literature correction factors for other seed models, the correction factors in this study diverge by up to 2.6% and 3.0% for iodine (with silver) and palladium respectively, indicating the need for seed-specific factors. Conclusion: The use of seed specific altitude correction factors can reduce uncertainty in the determination of air kerma strength. The empirical correction factors determined in this work can be applied in clinical quality assurance measurements of air kerma strength for two previously unpublished seed designs (IsoAid ADVANTAGE™ {sup 103}Pd and

  3. Verification of the calculation program for brachytherapy planning system of high dose rate (PLATO); Programa de verificacion del calculo para un sistema de planificacion de braquiterapia de alta tasa de dosis (PLATO)

    Energy Technology Data Exchange (ETDEWEB)

    Almansa, J.; Alaman, C.; Perez-Alija, J.; Herrero, C.; Real, R. del; Ososrio, J. L.

    2011-07-01

    In our treatments are performed brachytherapy high dose rate since 2007. The procedures performed include gynecological intracavitary treatment and interstitial. The treatments are performed with a source of Ir-192 activity between 5 and 10 Ci such that small variations in treatment times can cause damage to the patient. In addition the Royal Decree 1566/1998 on Quality Criteria in radiotherapy establishes the need to verify the monitor units or treatment time in radiotherapy and brachytherapy. All this justifies the existence of a redundant system for brachytherapy dose calculation that can reveal any abnormality is present.

  4. Locoregional recurrences following radical external beam irradiation and interstitial implantation for operable breast cancer - a twenty three year experience

    International Nuclear Information System (INIS)

    Leung, S.; Otmezguine, Y.; Calitchi, E.; Mazeron, J.J.; Le Bourgeois, J.P.; Pierquin, B.

    1986-01-01

    Locoregional recurrences are reported in 493 consecutive with T1 T2 N0 N1 breast cancer patients who were treated with radical external beam irradiation and interstitial 192 Ir implant between 1961 and 1979. Follow-up ranges from 5-23 years (mean 10 years) with 195 patients having 10-23 years follow up (mean 12 years). There were 51 (10%) locoregional recurrences with 34 mammary, 14 combined mammary/axillary and 3 isolated axillary recurrences. The 10 year relapse rate was 20/195 (10%). The results have been compared with other series in the literature and changes in the authors' current protocol are described. (Auth.)

  5. Carcinological results at five years of the prostate brachytherapy by iodine 125 implants. About 327 cases; Resultats carcinologiques a cinq ans de la curietherapie de prostate par implants d'iode 125. A propos de 327 cas

    Energy Technology Data Exchange (ETDEWEB)

    Peiffert, D.; Bernier, V.; Aletti, P.; Noela, A.; Marchesi, V. [Centre Alexis-Vautrin, 54 - Vandoeuvre-les-Nancy (France); Cormier, L.; Moreau, J.L. [CHU, Service d' Urologie, 54 - Nancy (France)

    2006-11-15

    The prostate brachytherapy by iodine 125 implants is reserved to patients damaged by a favourable prognosis cancer. The objective of this study is to present the carcinological results at five years in these indications with a technique keeping urethra. The achieving of a P.S.A. nadir inferior to 0.5 ng/ml is long. the technique used in this study keeping urethra and respecting the inclusion criteria recommended allows to get results as less equivalent as these ones of big series previously published with a low failure rate. (N.C.)

  6. Production of Pd 103 seed from Rh targets for brachytherapy

    International Nuclear Information System (INIS)

    Afarideh, H.; Ardaneh, K.; Sadeghi, M.

    2000-01-01

    The suitability of a given radionuclide for brachytherapy is determined by its half-life, the type of energy, and abundance (number per decay) of its emission. The half-life of a radionuclide must be long enough to permit shipping and implant preparation with an acceptable loss of source strength due to decay, but it must also be short enough to permit source sizes sufficiently small for the intended application. Pd-103 is a low energy photon emitter available for permanent interstitial implantation. Pd-103 has energy and safety characteristics similar to I-125, but its initial peripheral dose rate is approximately three times greater. This may provide improved control of rapidly proliferating tumours. Although Pd-103 has been used for various kinds of cancers, it is almost exclusively used for prostate cancer, the most common cancer, and the death rate from this cancer is the highest. There are two cyclotron production routes for Pd-103, Ag (p,xn) 103 Pd and Rh (p,n) 103 Pd. For a cyclotron with low energy (such as 30Mev that we have in Iran, Karaj, NRCAM) only Rh target can be used. The target material should be deposited on a special designed Cu substrate and the separation process should isolate the desired radionuclide from target material as well as Cu. Our work plan for production of Pd 103 in Karaj, Iran, is as follows: In the first year of the CRP we are going to complete the literature survey of Pd production and perform the relevant experiments as described later. In the second year of the CRP we will construct suitable hot cells for Pd production and also do research for development of Pd seeds. In the last year of the CRP we are going to finalise all the work done during the last two years and propose the automation system for routine production

  7. Prostate HDR brachytherapy catheter displacement between planning and treatment delivery

    International Nuclear Information System (INIS)

    Whitaker, May; Hruby, George; Lovett, Aimee; Patanjali, Nitya

    2011-01-01

    Background and purpose: HDR brachytherapy is used as a conformal boost for treating prostate cancer. Given the large doses delivered, it is critical that the volume treated matches that planned. Our outpatient protocol comprises two 9 Gy fractions, two weeks apart. We prospectively assessed catheter displacement between CT planning and treatment delivery. Materials and methods: Three fiducial markers and the catheters were implanted under transrectal ultrasound guidance. Metal marker wires were inserted into 4 reference catheters before CT; marker positions relative to each other and to the marker wires were measured from the CT scout. Measurements were repeated immediately prior to treatment delivery using pelvic X-ray with marker wires in the same reference catheters. Measurements from CT scout and film were compared. For displacements of 5 mm or more, indexer positions were adjusted prior to treatment delivery. Results: Results are based on 48 implants, in 25 patients. Median time from planning CT to treatment delivery was 254 min (range 81–367 min). Median catheter displacement was 7.5 mm (range −2.9–23.9 mm), 67% of implants had displacement of 5 mm or greater. Displacements were predominantly caudal. Conclusions: Catheter displacement can occur in the 1–3 h between the planning CT scan and treatment. It is recommended that departments performing HDR prostate brachytherapy verify catheter positions immediately prior to treatment delivery.

  8. Intestinal morphological effect of brachytherapy of low rate of dose, administrated in therapeutic form and its clinical manifestations in uterine cervix tumors

    International Nuclear Information System (INIS)

    Mendoza, Carmen; Contreras, Manuel

    2005-01-01

    Brachytherapy is effective to eradicate cancer in the cervix, in order to obtain the control of disease we use high dose with vesical and rectum toxicity. The objective is to investigate if brachytherapy by itself is the cause of intestinal damage, to know in addition if the intensity of the clinic manifestations is in direct relation to the given radiation dose and this gets worse when it is received in several applications. Hypothesis: The intensity of the radiation with brachytherapy of low rate of dose is proportional to the degree of clinical manifestations and morphologic damage of the intestine. A prospective analysis was made inpatients with cancer of cervix from september 2000 to june 2004. Each patient who enters to the department of brachytherapy of the hospital must be done laboratory examination that includes plaque and coagulation test before being accepted. We use the clinical card and a table in order to register data concerning teletherapy, implants of brachytherapy of low rate of dose, symptoms of intestinal toxicity and details of colonoscopia. Subsequent to the hospitable discharge the patient is sent to gastroenterology for clinical evaluation and to realize colonoscopia. From september 2000 to june 2004, 540 patients entered, 80 patients (15%) displayed intestinal manifestations, all received teletherapy and brachytherapy, nobody else received brachytherapy in exclusive form and only one patient (0.1%) received the total of the dose in 2 applications. The equipment of teletherapy Primus with energy of 6 and 18 Mv and implants of brachytherapy Manchester were used (70/55 patients). 79 (98%) patients received dose between 85-75 Gy in one single application, 58 (72%) received the total of the dose to the tumor, 21 (26%) in vaginal mucosa. Discussion: Brachytherapy is the cause of the damages in the intestinal mucosa. (The author)

  9. Definition of medical event is to be based on the total source strength for evaluation of permanent prostate brachytherapy: A report from the American Society for Radiation Oncology.

    Science.gov (United States)

    Nag, Subir; Demanes, D Jeffrey; Hagan, Michael; Rivard, Mark J; Thomadsen, Bruce R; Welsh, James S; Williamson, Jeffrey F

    2011-10-01

    The Nuclear Regulatory Commission deems it to be a medical event (ME) if the total dose delivered differs from the prescribed dose by 20% or more. A dose-based definition of ME is not appropriate for permanent prostate brachytherapy as it generates too many spurious MEs and thereby creates unnecessary apprehension in patients, and ties up regulatory bodies and the licensees in unnecessary and burdensome investigations. A more suitable definition of ME is required for permanent prostate brachytherapy. The American Society for Radiation Oncology (ASTRO) formed a working group of experienced clinicians to review the literature, assess the validity of current regulations, and make specific recommendations about the definition of an ME in permanent prostate brachytherapy. The working group found that the current definition of ME in §35.3045 as "the total dose delivered differs from the prescribed dose by 20 percent or more" was not suitable for permanent prostate brachytherapy since the prostate volume (and hence the resultant calculated prostate dose) is dependent on the timing of the imaging, the imaging modality used, the observer variability in prostate contouring, the planning margins used, inadequacies of brachytherapy treatment planning systems to calculate tissue doses, and seed migration within and outside the prostate. If a dose-based definition for permanent implants is applied strictly, many properly executed implants would be improperly classified as an ME leading to a detrimental effect on brachytherapy. The working group found that a source strength-based criterion, of >20% of source strength prescribed in the post-procedure written directive being implanted outside the planning target volume is more appropriate for defining ME in permanent prostate brachytherapy. ASTRO recommends that the definition of ME for permanent prostate brachytherapy should not be dose based but should be based upon the source strength (air-kerma strength) administered.

  10. Comparison of 60Cobalt and 192Iridium sources in high dose rate afterloading brachytherapy

    International Nuclear Information System (INIS)

    Richter, J.; Baier, K.; Flentje, M.

    2008-01-01

    Purpose: 60 Co sources with dimensions identical to those of 192 Ir have recently been made available in clinical brachytherapy. A longer half time reduces demands on logistics and quality assurance and perhaps costs. Material and Methods: Comparison of the physical properties of 60 Co and 192 Ir with regard to brachytherapy. Results: Required activities for the same air kerma rate are lower by a factor of 2.8 for 60 Co. Differential absorption in tissues of different densities can be neglected. Monte Carlo calculations demonstrate that integral dose due to radial dose fall off is higher for 192 Ir in comparison to 60 Co within the first 22 cm from the source (normalization at 1 cm). At larger distances this relationship is reversed. Conclusion: Clinical examples for intracavitary and interstitial applications however, show practically identical dose distributions in the treatment volume. (orig.)

  11. Selective tumor irradiation by infusional brachytherapy in nonresectable pancreatic cancer: a phase I study

    International Nuclear Information System (INIS)

    Order, Stanley E.; Siegel, Jeffry A.; Principato, Robert; Zeiger, Louis E.; Johnson, Elizabeth; Lang, Patricia; Lustig, Robert; Wallner, Paul E.

    1996-01-01

    Purpose: Selective high-dose radiation of solid tumors has been a goal of radiation oncology. The physiological barriers of solid tumors (high interstitial tumor pressure, reduced tumor vascularity, and poor perfusion) have been major barriers in achieving significant tumor dose of systemically infused radioconjugates. Direct tumor infusional brachytherapy overcomes these barriers and leads to selective high tumor doses. Methods and Materials: The development of interstitial tumor infusion of macroaggregated albumin (MAA) followed by colloidal chromic phosphate 32 P has overcome solid tumor obstacles in 47 patients with nonresectable pancreatic cancer in a Phase I dose escalation study. The colloidal 32 P infusion was followed by external radiation and five fluorouracil. Results: Of the 28 patients with cancer limited to the pancreas, 15 of 16 patients retained 86-100% (mean 96%) of the infused colloidal 32 P isotope. While the other 12 patients had partial shunting to the liver, shunting to the liver was due to high interstitial resistance with tumor dose deposition of 17-88% (mean 52%). Of the 19 patients with metastatic pancreas cancer, colloidal 32 P tumor deposition ranged from 22 to 100% of the infused dose (mean 79%). The less than optimal tumor deposition led to our increasing the MAA from 600,000 to 1.5-2.5 million particles. Interstitial dexamethasone 2 mg and later 4 mg was infused first and prevented liver shunting by somehow reducing tumor resistance. The median survival in 28 Phase I patients with nonresectable pancreas cancer without metastasis, was 12 months. No significant toxicity occurred when treatment was limited to two infusions with as much as 30 mCi each. The maximum tumor dose was 17,000 Gy (1.700,000 cGy). In 19 non-resectable pancreatic cancer patients with metastasis, a 6.9 months median survival was observed. Conclusions: Infusional brachytherapy is an outpatient procedure that delivers high-dose radiation selectively to pancreatic

  12. The source-skin distance measuring bridge: A method to avoid radiation teleangiectasia in the skin after interstitial therapy for breast cancer

    International Nuclear Information System (INIS)

    Van Limbergen, E.; Briot, E.; Drijkoningen, M.

    1990-01-01

    Inappropriate positioning of interstitial iridium 192 implants, used as booster dose in the breast conserving treatment of mammary cancer, may cause disturbing teleangiectasia of the breast skin, when high radiation doses are delivered on the dermal blood vessels. Based on the localization of the vascular plexuses in human breast skin, and on the dose distribution around different types of interstitial implants, a method is described to avoid overlap between the high dose area of the implant and the blood vessels in the skin. The latter are demonstrated to run within the first 5 mm under the epiderm. For source lengths varying from 5 to 8 cm, simple mathematical relations exist between the maximal security margin (MSM) and intersource distance (E) for single plane implants (MSM = 0.4 (E + 1)), double plane square implants (MSM = 0.4 E) and double plane triangular implants (MSM = 0.4 (E - 1)). We developed a device to measure precisely the distance between the radioactive wires and the overlying skin, along the whole source trajectory. Using this method, the occurrence of teleangiectasia in the breast skin after interstitial implants with Ir 192 may be significantly reduced

  13. Assessment of microseeds biodegradability of Sm and Sm:Ba splenic implants in rabbits

    International Nuclear Information System (INIS)

    Siqueira, Savio Lana; Barroso, Thiago Vinicius Villar; Campos, Tarcisio P.R.

    2009-01-01

    The radioactive interstitial implants have applications in controlling neoplasm in several regions of the human body. Currently the permanent brachytherapy seeds implanted in the spleen and other organs are made of I-125 seeds. After the total emission of radiation, the metal encapsulated seed remains inert in the implanted area. Seeds of bioactive ceramics have been prepared with Sm-152 incorporation to be activated in Sm-153. This study aimed to develop surgical technique for implanting biodegradable micro-seeds in the spleen of the rabbit. Three micro-seeds were introduced by hypodermic needle in the spleen in eight rabbits by median laparotomy. Subsequently, there were clinical and functional reactions of the animal to the implanted foreign body. The other objective was to perform the animal monitoring by radiography, produced in time sequence, and pathological studies of a fragment of the spleens of rabbits. The results show the effectiveness of surgery, the identification of the implanted material by radiography in vivo, and the biocompatibility of micro-seeds most of Sm and Sm:Ba. These seeds of reduced volume, 0.3x 1.6 mm, could be monitored for radiological studies in 2 periods: early and later implant. On the later studies, radiography was taken at 60d post-implant. Biopsies were taken and radiographs of the samples were also performed for evidencing the degradation state of the seeds. The results of the two groups of four rabbits are presented. They show partial degradation of the seed verified by radiographic contrast which is related to the atomic number of the elements and mass density in the seed. The biopsy showed that the ceramic is clearly absorbed by the spleen tissue and form tissue-implant interface. The histological slides showed an inflammatory reaction with presence of fibrosis of the giant cell foreign body. In conclusion, the radiograph shows a suitable noninvasive technique for monitoring the degradation of micro-seed ceramics in vivo

  14. Assessment of microseeds biodegradability of Sm and Sm:Ba splenic implants in rabbits

    Energy Technology Data Exchange (ETDEWEB)

    Siqueira, Savio Lana; Barroso, Thiago Vinicius Villar [Faculdade de Ciencias Medicas de Minas Gerais, Belo Horizonte, MG (Brazil). Dept. de Anatomia; Campos, Tarcisio P.R., E-mail: campos@nuclear.ufmg.b [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Programa de Ciencias e Tecnicas Nucleares

    2009-07-01

    The radioactive interstitial implants have applications in controlling neoplasm in several regions of the human body. Currently the permanent brachytherapy seeds implanted in the spleen and other organs are made of I-125 seeds. After the total emission of radiation, the metal encapsulated seed remains inert in the implanted area. Seeds of bioactive ceramics have been prepared with Sm-152 incorporation to be activated in Sm-153. This study aimed to develop surgical technique for implanting biodegradable micro-seeds in the spleen of the rabbit. Three micro-seeds were introduced by hypodermic needle in the spleen in eight rabbits by median laparotomy. Subsequently, there were clinical and functional reactions of the animal to the implanted foreign body. The other objective was to perform the animal monitoring by radiography, produced in time sequence, and pathological studies of a fragment of the spleens of rabbits. The results show the effectiveness of surgery, the identification of the implanted material by radiography in vivo, and the biocompatibility of micro-seeds most of Sm and Sm:Ba. These seeds of reduced volume, 0.3x 1.6 mm, could be monitored for radiological studies in 2 periods: early and later implant. On the later studies, radiography was taken at 60d post-implant. Biopsies were taken and radiographs of the samples were also performed for evidencing the degradation state of the seeds. The results of the two groups of four rabbits are presented. They show partial degradation of the seed verified by radiographic contrast which is related to the atomic number of the elements and mass density in the seed. The biopsy showed that the ceramic is clearly absorbed by the spleen tissue and form tissue-implant interface. The histological slides showed an inflammatory reaction with presence of fibrosis of the giant cell foreign body. In conclusion, the radiograph shows a suitable noninvasive technique for monitoring the degradation of micro-seed ceramics in vivo

  15. Optimization in brachytherapy with the implementation of Radiobiology

    International Nuclear Information System (INIS)

    Duran, M.P.; Bourel, V.J.; Rodriguez, I.; Torre, M. de la; Caneva, S.

    1998-01-01

    In the brachytherapy planning treatments with High dose rates (HDR), the optimization algorithms used are based in dosimetric considerations and/or geometric ones, ignoring the radiobiological response of the tissue treated. In this work we wish to show the implementation of radiobiological concepts in the optimization. Assuming that the subtiles differences that result in the dose distribution among the different optimization models which are not visible in an isodose plane, it is studied how is classically make it , the quality implant through natural histograms about dose volumes and the resulting parameters. Also is studied the necrosis probability which may be caused by the choice of some optimization model, allowing with this the choice of the best implant. (Author)

  16. Automation of brachytherapy planning based on RADPLAN system

    International Nuclear Information System (INIS)

    Costa, Helder Rodrigues da; Campos, Tarcisio P.R.

    2000-01-01

    RADPLAN is a software to automate the clinical planning of doses in brachytherapy managing the radionuclide source position and time. It intends to be a important tools to assist cancer treatment in the radiotherapy services. It evaluates the tridimensional isodoses for a set of radioactive sources implanted on tissue or placed inside a body cavity. RADPLAN allows visualization of virtual frames in which isodose curves overlap medical images produced in computerized tomography), X-ray and nuclear magnetic resonance equipment. In this work, the software function is explained and a hypothetic case of medical information management is presented, specifically illustrations of isodoses curves obtained for a spatial distribution of Ir 192 wires implanted in a breast. (author)

  17. 3-D conformal HDR brachytherapy as monotherapy for localized prostate cancer. A pilot study

    International Nuclear Information System (INIS)

    Martin, T.; Baltas, D.; Kurek, R.; Roeddiger, S.; Kontova, M.; Anagnostopoulos, G.; Skazikis, G.; Zamboglou, N.; Dannenberg, T.; Buhleier, T.; Tunn, U.

    2004-01-01

    Purpose: pilot study to evaluate feasibility, acute toxicity and conformal quality of three-dimensional (3-D) conformal high-dose-rate (HDR) brachytherapy as monotherapy for localized prostate cancer using intraoperative real-time planning. Patients and methods: between 05/2002 and 05/2003, 52 patients with prostate cancer, prostate-specific antigen (PSA) ≤ 10 ng/ml, Gleason score ≤ 7 and clinical stage ≤ T2a were treated. Median PSA was 6.4 ng/ml and median Gleason score 5. 24/52 patients had stage T1c and 28/52 stage T2a. For transrectal ultrasound-(TRUS-)guided transperineal implantation of flexible plastic needles into the prostate, the real-time HDR planning system SWIFT trademark was used. After implantation, CT-based 3-D postplanning was performed. All patients received one implant for four fractions of HDR brachytherapy in 48 h using a reference dose (D ref ) of 9.5 Gy to a total dose of 38.0 Gy. Dose-volume histograms (DVHs) were analyzed to evaluate the conformal quality of each implant using D 90 , D 10 urethra, and D 10 rectum. Acute toxicity was evaluated using the CTC (common toxicity criteria) scales. Results: median D 90 was 106% of D ref (range: 93-115%), median D 10 urethra 159% of D ref (range: 127-192%), and median D 10 rectum 55% of D ref (range: 35-68%). Median follow-up is currently 8 months. In 2/52 patients acute grade 3 genitourinary toxicity was observed. No gastrointestinal toxicity > grade 1 occurred. Conclusion: 3-D conformal HDR brachytherapy as monotherapy using intraoperative real-time planning is a feasible and highly conformal treatment for localized prostate cancer associated with minimal acute toxicity. Longer follow-up is needed to evaluate late toxicity and biochemical control. (orig.)

  18. Hydrogen interstitial in H-ion implanted ZnO bulk single crystals: Evaluation by elastic recoil detection analysis and electron paramagnetic resonance

    Energy Technology Data Exchange (ETDEWEB)

    Kaida, T.; Kamioka, K.; Nishimura, T. [College of Engineering and Research Center of Ion Beam Technology, Hosei University, Koganei, Tokyo 184-8584 (Japan); Kuriyama, K., E-mail: kuri@ionbeam.hosei.ac.jp [College of Engineering and Research Center of Ion Beam Technology, Hosei University, Koganei, Tokyo 184-8584 (Japan); Kushida, K. [Department of Arts and Sciences, Osaka Kyoiku University, Kashiwara, Osaka 582-8582 (Japan); Kinomura, A. [National Institute of Advanced Industrial Science and Technology, Tsukuba, Ibaraki 305-8568 (Japan)

    2015-12-15

    The origins of low resistivity in H ion-implanted ZnO bulk single crystals are evaluated by elastic recoil detection analysis (ERDA), electron paramagnetic resonance (EPR), and Van der Pauw methods. The H-ion implantation (peak concentration: 5.0 × 10{sup 15} cm{sup −2}) into ZnO is performed using a 500 keV implanter. The maximum of the concentration of the implanted H estimated by a TRIM simulation is at 3600 nm in depth. The resistivity decreases from ∼10{sup 3} Ω cm for un implanted ZnO to 6.5 Ω cm for as-implanted, 2.3 × 10{sup −1} Ω cm for 200 °C annealed, and 3.2 × 10{sup −1} Ω cm for 400 °C annealed samples. The ERDA measurements can evaluate the concentration of hydrogens which move to the vicinity of the surface (surface to 300 nm or 100 nm) because of the diffusion by the annealing at 200 °C and 400 °C. The hydrogen concentration near the surface estimated using the 2.0 MeV helium beam is ∼3.8 × 10{sup 13} cm{sup −2} for annealed samples. From EPR measurements, the oxygen vacancy of +charge state (V{sub o}{sup +}) is observed in as-implanted samples. The V{sub o}{sup +} related signal (g = 1.96) observed under no illumination disappears after successive illumination with a red LED and appears again with a blue light illumination. The activation energy of as-implanted, 200 °C annealed, and 400 °C annealed samples estimated from the temperature dependence of carrier concentration lies between 29 meV and 23 meV, suggesting the existence of H interstitial as a shallow donor level.

  19. Interobserver variations of target volume delineation and its impact on irradiated volume in accelerated partial breast irradiation with intraoperative interstitial breast implant

    Directory of Open Access Journals (Sweden)

    Ritu Raj Upreti

    2017-02-01

    Full Text Available Purpose: To investigate the interobserver variations in delineation of lumpectomy cavity (LC and clinical target volume (CTV, and its impact on irradiated volume in accelerated partial breast irradiation using intraoperative multicatheter brachytherapy. Material and methods : Delineation of LC and CTV was done by five radiation oncologists on planning computed tomography (CT scans of 20 patients with intraoperative interstitial breast implant. Cavity visualization index (CVI, four-point index ranging from (0 = poor to (3 = excellent was created and assigned by observers for each patient. In total, 200 contours for all observers and 100 treatment plans were evaluated. Spatial concordance (conformity index, CI common , and CIgen, average shift in the center of mass (COM, and ratio of maximum and minimum volumes (V max /V min of LC and CTV were quantified among all observers and statistically analyzed. Variation in active dwell positions (0.5 cm step for each catheter, total reference air kerma (TRAK, volume enclosed by prescription isodose (V100% among observers and its spatial concordance were analyzed. Results : The mean ± SD CI common of LC and CTV was 0.54 ± 0.09, and 0.58 ± 0.08, respectively. Conformity index tends to increase, shift in COM and V max /V min decrease significantly (p < 0.05, as CVI increased. Out of total 309 catheters, 29.8% catheters had no change, 29.8% and 17.5% catheters had variations of 1 and 2 dwell positions (0.5 cm and 1 cm, respectively. 9.3% catheters shown variations ≥ 10 dwell positions (5 cm. The mean ± SD CI common of V100% was 0.75 ± 0.11. The mean observed V max /V min of prescription isodose and TRAK was 1.18 (range, 1.03 to 1.56 and 1.11 (range, 1.03 to 1.35, respectively. Conclusions : Interobserver variability in delineation of target volume was found to be significantly related to CVI. Smaller variability was observed with excellent visualization of LC. Interobserver variations showed dosimetric

  20. DNA-PKcs Expression Is a Predictor of Biochemical Recurrence After Permanent Iodine 125 Interstitial Brachytherapy for Prostate Cancer

    International Nuclear Information System (INIS)

    Molina, Sarah; Guerif, Stéphane; Garcia, Alexandre; Debiais, Céline; Irani, Jacques; Fromont, Gaëlle

    2016-01-01

    Purpose: Predictive factors for biochemical recurrence (BCR) in localized prostate cancer (PCa) after brachytherapy are insufficient to date. Cellular radiosensitivity depends on DNA double-strand breaks, mainly repaired by the nonhomologous end-joining (NHEJ) system. We analyzed whether the expression of NHEJ proteins can predict BCR in patients treated by brachytherapy for localized PCa. Methods and Materials: From 983 PCa cases treated by brachytherapy between March 2000 and March 2012, 167 patients with available biopsy material suitable for in situ analysis were included in the study. The median follow-up time was 47 months. Twenty-nine patients experienced BCR. All slides were reviewed to reassess the Gleason score. Expression of the key NHEJ proteins DNA-PKcs, Ku70, and Ku80, and the proliferation marker Ki67, was studied by immunohistochemistry performed on tissue microarrays. Results: The Gleason scores after review (P=.06) tended to be associated with BCR when compared with the score initially reported (P=.74). Both the clinical stage (P=.02) and the pretreatment prostate-specific antigen level (P=.01) were associated with biochemical failure. Whereas the expression of Ku80 and Ki67 were not predictive of relapse, positive DNA-PKcs nuclear staining (P=.003) and higher Ku70 expression (P=.05) were associated with BCR. On multivariate analysis, among pretreatment variables, only DNA-PKcs (P=.03) and clinical stage (P=.02) remained predictive of recurrence. None of the patients without palpable PCa and negative DNA-PKcs expression experienced biochemical failure, compared with 32% of men with palpable and positive DNA-PKcs staining that recurred. Conclusions: Our results suggest that DNA-PKcs could be a predictive marker of BCR after brachytherapy, and this might be a useful tool for optimizing the choice of treatment in low-risk PCa patients.

  1. DNA-PKcs Expression Is a Predictor of Biochemical Recurrence After Permanent Iodine 125 Interstitial Brachytherapy for Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Molina, Sarah [Department of Pathology, INSERM UMR1069, CHU/Université de Tours, Tours (France); Department of Radiation Oncology, CHU/Université de Poitiers, Poitiers (France); Guerif, Stéphane; Garcia, Alexandre [Department of Radiation Oncology, CHU/Université de Poitiers, Poitiers (France); Debiais, Céline [Department of Pathology, CHU/Université de Poitiers, Poitiers (France); Irani, Jacques [Department of Urology, CHU/Université de Poitiers, Poitiers (France); Fromont, Gaëlle, E-mail: gaelle.fromont-hankard@univ-tours.fr [Department of Pathology, INSERM UMR1069, CHU/Université de Tours, Tours (France)

    2016-07-01

    Purpose: Predictive factors for biochemical recurrence (BCR) in localized prostate cancer (PCa) after brachytherapy are insufficient to date. Cellular radiosensitivity depends on DNA double-strand breaks, mainly repaired by the nonhomologous end-joining (NHEJ) system. We analyzed whether the expression of NHEJ proteins can predict BCR in patients treated by brachytherapy for localized PCa. Methods and Materials: From 983 PCa cases treated by brachytherapy between March 2000 and March 2012, 167 patients with available biopsy material suitable for in situ analysis were included in the study. The median follow-up time was 47 months. Twenty-nine patients experienced BCR. All slides were reviewed to reassess the Gleason score. Expression of the key NHEJ proteins DNA-PKcs, Ku70, and Ku80, and the proliferation marker Ki67, was studied by immunohistochemistry performed on tissue microarrays. Results: The Gleason scores after review (P=.06) tended to be associated with BCR when compared with the score initially reported (P=.74). Both the clinical stage (P=.02) and the pretreatment prostate-specific antigen level (P=.01) were associated with biochemical failure. Whereas the expression of Ku80 and Ki67 were not predictive of relapse, positive DNA-PKcs nuclear staining (P=.003) and higher Ku70 expression (P=.05) were associated with BCR. On multivariate analysis, among pretreatment variables, only DNA-PKcs (P=.03) and clinical stage (P=.02) remained predictive of recurrence. None of the patients without palpable PCa and negative DNA-PKcs expression experienced biochemical failure, compared with 32% of men with palpable and positive DNA-PKcs staining that recurred. Conclusions: Our results suggest that DNA-PKcs could be a predictive marker of BCR after brachytherapy, and this might be a useful tool for optimizing the choice of treatment in low-risk PCa patients.

  2. Three-dimensional tomosynthetic image restoration for brachytherapy source localization

    International Nuclear Information System (INIS)

    Persons, Timothy M.

    2001-01-01

    Tomosynthetic image reconstruction allows for the production of a virtually infinite number of slices from a finite number of projection views of a subject. If the reconstructed image volume is viewed in toto, and the three-dimensional (3D) impulse response is accurately known, then it is possible to solve the inverse problem (deconvolution) using canonical image restoration methods (such as Wiener filtering or solution by conjugate gradient least squares iteration) by extension to three dimensions in either the spatial or the frequency domains. This dissertation presents modified direct and iterative restoration methods for solving the inverse tomosynthetic imaging problem in 3D. The significant blur artifact that is common to tomosynthetic reconstructions is deconvolved by solving for the entire 3D image at once. The 3D impulse response is computed analytically using a fiducial reference schema as realized in a robust, self-calibrating solution to generalized tomosynthesis. 3D modulation transfer function analysis is used to characterize the tomosynthetic resolution of the 3D reconstructions. The relevant clinical application of these methods is 3D imaging for brachytherapy source localization. Conventional localization schemes for brachytherapy implants using orthogonal or stereoscopic projection radiographs suffer from scaling distortions and poor visibility of implanted seeds, resulting in compromised source tracking (reported errors: 2-4 mm) and dosimetric inaccuracy. 3D image reconstruction (using a well-chosen projection sampling scheme) and restoration of a prostate brachytherapy phantom is used for testing. The approaches presented in this work localize source centroids with submillimeter error in two Cartesian dimensions and just over one millimeter error in the third

  3. Brachytherapy in vulvar cancer: analysis of 18 patients

    International Nuclear Information System (INIS)

    Frezza, G.; Baldissera, A.; Bernardi, L.; Bunkheila, F.; Galuppi, A.; Salvi, F.

    1996-01-01

    INTRODUCTION: Vulvar cancer is a rather common neoplasm in elderly patients. Surgery, followed eventually by postoperative radiotherapy, is the treatment of choice. The results of exclusive radiotherapy (external beam irradiation and/or brachytherapy) are not well defined and in the recent literature only small series are reported. Radiotherapy however is the only therapeutic option in patients who are not fit for radical surgery. It is thus necessary to review its indications and its modalities. PATIENTS METHODS AND RESULTS: From 1990 to 1994 18 pts with a diagnosis of squamous cell carcinoma of the vulva have been submitted to brachytherapy. Age ranged from 60 to 92 years (mean age 76, 1 ys). 14 pts were treated at diagnosis (11 pts) or for recurrent disease after surgery (3 pts). In 8 of them brachytherapy (total dose 35-45 Gy, dose rate: 0,4-0,78 Gy/h) was preceded by external beam irradiation (Co60 or electron beam, 40-50 Gy to primary and inguinal nodes); 6 pts were treated with brachytherapy alone (58-60 Gy; dose rate 0,44-0,63 Gy/h). 4 pts underwent to brachytherapy alone for local recurrence after surgery and postoperative radiotherapy (total dose 45-60 Gy; dose rate 0,37-0,49 Gy/h). Brachytherapy was always performed with 192 Ir. Plastic tubes (2 to 5 lines) were used for single plane implantation of small exophytic lesions limited to the labia (8 cases); a perineal template (10 cases) was employed in lesions extended to the vaginal mucosa or involving the clitoris or the area of the perineum. (10(14)) pts treated at diagnosis are alive and free from local recurrence after 11-48 mos. 3 of them, treated with brachytherapy alone, have presented a nodal recurrence in the groin after 14, 15 and 27 mos. respectively. All of them are alive and free from disease after surgery and external radiotherapy. None of the pts treated for recurrent disease after surgery + external beam radiotherapy has achieved a local control. CONCLUSION: Brachytherapy alone or

  4. Interstitial hyperthermia and iridium-192 treatment alone vs. interstitial iridium-192 treatment/hyperthermia and low dose cisplatinum infusion in the treatment of locally advanced head and neck malignancies

    International Nuclear Information System (INIS)

    Schreiber, David P.; Overett, Thomas K.

    1995-01-01

    Purpose: To determine whether the addition of low dose platinum infusional chemotherapy adds to the effectiveness of interstitial hyperthermia/iridium-192 management of locally advanced head and neck malignancies. Methods and Materials: From 1987 to 1993, 36 patients with locally advanced head and neck malignancies were treated locally with interstitial hyperthermia and iridium-192 as part or all of their management. Twenty-two of the above-mentioned patients also received low dose infusional cisplatinum chemotherapy at 20 mg/M 2 per day during the time of the implant. No patient received greater than 100 mg/M 2 total dose. Implant times ranged from 38.5 to 134 h and total doses delivered with the radiation implants ranged from 15 to 39.9 Gy. Average implant volume was 50 cc. Twenty-three patients received external beam irradiation supplementation in a dose range from 25.2 to 64 Gy. Results: Median follow-up for the entire group was 8, months with 7 months for the chemotherapy group vs. 12 months for the no-chemotherapy group. Freedom from relapse rates for the chemotherapy group vs. the no-chemotherapy group were 70% at 41 months vs. 63% at 60 months, p not significant (p = NS). Overall survival by Life Table Analysis was 28% for the chemotherapy group at 41 months vs. 31% for the no-chemotherapy group at 60 months (p = NS). Complete response (CR) rates were 93% for the chemotherapy group vs. 86% for the no-chemotherapy group. Seven patients in the chemotherapy group had recurrent disease and four patients in the no-chemotherapy group were being treated for recurrent disease. Complication rates were similar in both groups, with two patients in the chemotherapy arm requiring hyperbaric oxygen treatments and one patient in the no-chemotherapy arm requiring hyperbaric oxygen treatments (for soft tissue necrosis). Conclusion: It appears that low dose platinum infusional chemotherapy can be added safely to patients receiving interstitial iridium-192 implants along with

  5. Orgasm after prostate curietherapy with iodine 125 permanent implants for localized cancer of the prostate

    International Nuclear Information System (INIS)

    Delaunay, B.; Plante, P.; Huyghe, E.; Delannes, M.; Bachaud, J.-M.; Salloum, A.; Thoulouzan, M.; Soulie, M.; Delavierre, D.; Wagner, F.; Jonca, F.

    2011-01-01

    Orgasm is a domain of male sexuality that remains underreported in literature. Our aim was to realize the first detailed analysis of orgasm in patients treated by 125 I permanent prostate brachytherapy for localized prostate cancer. In a series of 270 sexually active men treated by prostate brachytherapy ( 125 I permanent implantation), 241 (89%), mean age of 65 (43 80), participated in a mailed survey about sexual function after a mean time of 36 months (9 70). Erectile and ejaculatory functions and orgasm were explored using a mailed questionnaire. Two questions focused on orgasm. The first was about quality of orgasm (fast/intense/late, difficult/weak/absent) and the second about the presence of painful orgasm and its frequency (always/sometimes/often). After prostate brachytherapy, 81.3% of sexually active men conserved ejaculation and 90% orgasm. There was a significant deterioration of the quality of orgasm (P ≡ 0.0001). More than 50% of the patients had an altered orgasm (weak, difficult, absent) after brachytherapy, vs 16% before implantation (P ≡ 0.001). Men with a diminished ejaculation volume often had a weak/difficult orgasm (P ≡ 0.007). Neoadjuvant hormonal therapy did not seem to impact the quality of orgasm or the frequency of painful ejaculation. Patients who had an IIEF-5 score higher than 12 had frequently intense orgasm (26.7% vs 2.7%; P < 0.001) after brachytherapy. Sixty patients (30.3%) experienced often/sometimes painful ejaculation 12.9% (n ≡ 31) before implantation (P ≡ 0.0001). Most of the patients treated by prostate brachytherapy conserved orgasm after treatment. However, most of the patients described a deterioration of the quality of orgasm. (authors)

  6. Urethral dose and increment of international prostate symptom score (IPSS) in transperineal permanent interstitial implant (TPI) of prostate cancer

    International Nuclear Information System (INIS)

    Murakami, N.; Itami, J.; Okuma, K.; Marino, H.; Ban, T.; Nakazato, M.; Kanai, K.; Naoi, K.; Fuse, M.; Nakagawa, K.

    2008-01-01

    Purpose: to find the factors which influence the acute increment of international prostate symptom score (IPSS) after transperineal permanent interstitial implant (TPI) using 125 I seeds. Patients and methods: from April 2004 through September 2006, 104 patients with nonmetastatic prostate cancer underwent TPI without external-beam irradiation. Median patient age was 70 years with a median follow-up of 13.0 months. 73 patients (70%) received neoadjuvant hormone therapy. The increment of IPSS was defined as the difference between pre- and postimplant maximal IPSS. Clinical, treatment, and dosimetric parameters evaluated included age, initial prostate-specific antigen, Gleason Score, neoadjuvant hormone therapy, initial IPSS, post-TPI prostatic volume, number of implanted seeds, prostate V 100 , V 150 , D 90 , urethral D max , and urethral D 90 . In order to further evaluate detailed urethral doses, the base and apical urethra were defined and the dosimetric parameters were calculated. Results: the IPSS peaked 3 months after TPI and returned to baseline at 12-15 months. Multivariate analysis demonstrated a statistically significant correlation of post-TPI prostatic volume, number of implanted seeds, and the dosimetric parameters of the base urethra with IPSS increment. Conclusion: the base urethra appears to be susceptible to radiation and the increased dose to this region deteriorates IPSS. It remains unclear whether the base urethral dose relates to the incidence of late urinary morbidities. (orig.)

  7. Dwell time modulation restrictions do not necessarily improve treatment plan quality for prostate HDR brachytherapy

    International Nuclear Information System (INIS)

    Balvert, Marleen; Gorissen, Bram L; Den Hertog, Dick; Hoffmann, Aswin L

    2015-01-01

    Inverse planning algorithms for dwell time optimisation in interstitial high-dose-rate (HDR) brachytherapy may produce solutions with large dwell time variations within catheters, which may result in undesirable selective high-dose subvolumes. Extending the dwell time optimisation model with a dwell time modulation restriction (DTMR) that limits dwell time differences between neighboring dwell positions has been suggested to eliminate this problem. DTMRs may additionally reduce the sensitivity for uncertainties in dwell positions that inevitably result from catheter reconstruction errors and afterloader source positioning inaccuracies. This study quantifies the reduction of high-dose subvolumes and the robustness against these uncertainties by applying a DTMR to template-based prostate HDR brachytherapy implants. Three different DTMRs were consecutively applied to a linear dose-based penalty model (LD) and a dose-volume based model (LDV), both obtained from literature. The models were solved with DTMR levels ranging from no restriction to uniform dwell times within catheters in discrete steps. Uncertainties were simulated on clinical cases using in-house developed software, and dose-volume metrics were calculated in each simulation. For the assessment of high-dose subvolumes, the dose homogeneity index (DHI) and the contiguous dose volume histogram were analysed. Robustness was measured by the improvement of the lowest D 90% of the planning target volume (PTV) observed in the simulations. For (LD), a DTMR yields an increase in DHI of approximately 30% and reduces the size of the largest high-dose volume by 2–5 cc. However, this comes at a cost of a reduction in D 90% of the PTV of 10%, which often implies that it drops below the desired minimum of 100%. For (LDV), none of the DTMRs were able to improve high-dose volume measures. DTMRs were not capable of improving robustness of PTV D 90% against uncertainty in dwell positions for both models. (paper)

  8. Iridium-192 implants in the treatment of tonsillar region malignancies

    International Nuclear Information System (INIS)

    Puthawala, A.A.; Syed, A.M.; Gates, T.C.

    1985-01-01

    A total of 127 patients with histologically proved diagnosis of carcinoma of the tonsillar region and soft palate were treated over the past ten years utilizing interstitial iridium-192 implants. Eighty patients were treated for primary tumors and 47 patients for either recurrent or persistent tumors after definitive irradiation and/or surgery. All patients with primary tumors were treated by a combination of external megavoltage irradiation and interstitial iridium-192 implants, whereas patients with recurrent tumors were treated by interstitial implants alone. In the primary group, 95% of patients with T1 and T2 lesions and 79% with T3 and T4 lesions achieved complete local tumor control. The three-year absolute disease-free survival rate was 72%. Seventy-five percent local tumor control was obtained in patients with recurrent disease, with two-year absolute disease-free survival of 42%. Treatment-related complications such as soft-tissue necrosis or osteoradionecrosis occurred in 6% of patients in the primary group and 23% in recurrent group. This treatment regimen offers an excellent locoregional control with no significant functional or esthetic impairment. Most patients with primary lesions who fail this regimen can be salvaged by surgery or reirradiation using interstitial implants

  9. Oxidation-enhanced diffusion of boron in very low-energy N2+-implanted silicon

    Science.gov (United States)

    Skarlatos, D.; Tsamis, C.; Perego, M.; Fanciulli, M.

    2005-06-01

    In this article we study the interstitial injection during oxidation of very low-energy nitrogen-implanted silicon. Buried boron δ layers are used to monitor the interstitial supersaturation during the oxidation of nitrogen-implanted silicon. No difference in boron diffusivity enhancement was observed compared to dry oxidation of nonimplanted samples. This result is different from our experience from N2O oxynitridation study, during which a boron diffusivity enhancement of the order of 20% was observed, revealing the influence of interfacial nitrogen on interstitial kinetics. A possible explanation may be that implanted nitrogen acts as an excess interstitial sink in order to diffuse towards the surface via a non-Fickian mechanism. This work completes a wide study of oxidation of very low-energy nitrogen-implanted silicon related phenomena we performed within the last two years [D. Skarlatos, C. Tsamis, and D. Tsoukalas, J. Appl. Phys. 93, 1832 (2003); D. Skarlatos, E. Kapetanakis, P. Normand, C. Tsamis, M. Perego, S. Ferrari, M. Fanciulli, and D. Tsoukalas, J. Appl. Phys. 96, 300 (2004)].

  10. Local anesthesia for prostate brachytherapy

    International Nuclear Information System (INIS)

    Wallner, Kent; Simpson, Colleen; Roof, James; Arthurs, Sandy; Korssjoen, Tammy; Sutlief, Steven

    1999-01-01

    Purpose: To demonstrate the technique and feasibility of prostate brachytherapy performed with local anesthesia only. Methods and Materials: A 5 by 5 cm patch of perineal skin and subcutaneous tissue is anesthetized by local infiltration of 10 cc of 1% lidocaine with epinephrine, using a 25-gauge 5/8-inch needle. Immediately following injection into the subcutaneous tissues, the deeper tissues, including the pelvic floor and prostate apex, are anesthetized by injecting 15 cc lidocaine solution with approximately 8 passes of a 20-gauge 1.0-inch needle. Following subcutaneous and peri-apical lidocaine injections, the patient is brought to the simulator suite and placed in leg stirrups. The transrectal ultrasound (TRUS) probe is positioned to reproduce the planning images and a 3.5- or 6.0-inch, 22-gauge spinal needle is inserted into the peripheral planned needle tracks, monitored by TRUS. When the tips of the needles reach the prostatic base, about 1 cc of lidocaine solution is injected in the intraprostatic track, as the needle is slowly withdrawn, for a total volume of 15 cc. The implants are done with a Mick Applicator, inserting and loading groups of two to four needles, so that a maximum of only about four needles are in the patient at any one time. During the implant procedure, an additional 1 cc of lidocaine solution is injected into one or more needle tracks if the patient experiences substantial discomfort. The total dose of lidocaine is generally limited to 500 mg (50 ml of 1% solution). Results: To date, we have implanted approximately 50 patients in our simulator suite, using local anesthesia. Patients' heart rate and diastolic blood pressure usually showed moderate changes, consistent with some discomfort. The time from first subcutaneous injection and completion of the source insertion ranged from 35 to 90 minutes. Serum lidocaine levels were below or at the low range of therapeutic. There has been only one instance of acute urinary retention in the

  11. GGEMS-Brachy: GPU GEant4-based Monte Carlo simulation for brachytherapy applications

    Science.gov (United States)

    Lemaréchal, Yannick; Bert, Julien; Falconnet, Claire; Després, Philippe; Valeri, Antoine; Schick, Ulrike; Pradier, Olivier; Garcia, Marie-Paule; Boussion, Nicolas; Visvikis, Dimitris

    2015-07-01

    In brachytherapy, plans are routinely calculated using the AAPM TG43 formalism which considers the patient as a simple water object. An accurate modeling of the physical processes considering patient heterogeneity using Monte Carlo simulation (MCS) methods is currently too time-consuming and computationally demanding to be routinely used. In this work we implemented and evaluated an accurate and fast MCS on Graphics Processing Units (GPU) for brachytherapy low dose rate (LDR) applications. A previously proposed Geant4 based MCS framework implemented on GPU (GGEMS) was extended to include a hybrid GPU navigator, allowing navigation within voxelized patient specific images and analytically modeled 125I seeds used in LDR brachytherapy. In addition, dose scoring based on track length estimator including uncertainty calculations was incorporated. The implemented GGEMS-brachy platform was validated using a comparison with Geant4 simulations and reference datasets. Finally, a comparative dosimetry study based on the current clinical standard (TG43) and the proposed platform was performed on twelve prostate cancer patients undergoing LDR brachytherapy. Considering patient 3D CT volumes of 400  × 250  × 65 voxels and an average of 58 implanted seeds, the mean patient dosimetry study run time for a 2% dose uncertainty was 9.35 s (≈500 ms 10-6 simulated particles) and 2.5 s when using one and four GPUs, respectively. The performance of the proposed GGEMS-brachy platform allows envisaging the use of Monte Carlo simulation based dosimetry studies in brachytherapy compatible with clinical practice. Although the proposed platform was evaluated for prostate cancer, it is equally applicable to other LDR brachytherapy clinical applications. Future extensions will allow its application in high dose rate brachytherapy applications.

  12. GGEMS-Brachy: GPU GEant4-based Monte Carlo simulation for brachytherapy applications

    International Nuclear Information System (INIS)

    Lemaréchal, Yannick; Bert, Julien; Schick, Ulrike; Pradier, Olivier; Garcia, Marie-Paule; Boussion, Nicolas; Visvikis, Dimitris; Falconnet, Claire; Després, Philippe; Valeri, Antoine

    2015-01-01

    In brachytherapy, plans are routinely calculated using the AAPM TG43 formalism which considers the patient as a simple water object. An accurate modeling of the physical processes considering patient heterogeneity using Monte Carlo simulation (MCS) methods is currently too time-consuming and computationally demanding to be routinely used. In this work we implemented and evaluated an accurate and fast MCS on Graphics Processing Units (GPU) for brachytherapy low dose rate (LDR) applications. A previously proposed Geant4 based MCS framework implemented on GPU (GGEMS) was extended to include a hybrid GPU navigator, allowing navigation within voxelized patient specific images and analytically modeled 125 I seeds used in LDR brachytherapy. In addition, dose scoring based on track length estimator including uncertainty calculations was incorporated. The implemented GGEMS-brachy platform was validated using a comparison with Geant4 simulations and reference datasets. Finally, a comparative dosimetry study based on the current clinical standard (TG43) and the proposed platform was performed on twelve prostate cancer patients undergoing LDR brachytherapy. Considering patient 3D CT volumes of 400  × 250  × 65 voxels and an average of 58 implanted seeds, the mean patient dosimetry study run time for a 2% dose uncertainty was 9.35 s (≈500 ms 10 −6 simulated particles) and 2.5 s when using one and four GPUs, respectively. The performance of the proposed GGEMS-brachy platform allows envisaging the use of Monte Carlo simulation based dosimetry studies in brachytherapy compatible with clinical practice. Although the proposed platform was evaluated for prostate cancer, it is equally applicable to other LDR brachytherapy clinical applications. Future extensions will allow its application in high dose rate brachytherapy applications. (paper)

  13. American Brachytherapy Society Task Group Report: Combination of brachytherapy and external beam radiation for high-risk prostate cancer.

    Science.gov (United States)

    Spratt, Daniel E; Soni, Payal D; McLaughlin, Patrick W; Merrick, Gregory S; Stock, Richard G; Blasko, John C; Zelefsky, Michael J

    To review outcomes for high-risk prostate cancer treated with combined modality radiation therapy (CMRT) utilizing external beam radiation therapy (EBRT) with a brachytherapy boost. The available literature for high-risk prostate cancer treated with combined modality radiation therapy was reviewed and summarized. At this time, the literature suggests that the majority of high-risk cancers are curable with multimodal treatment. Several large retrospective studies and three prospective randomized trials comparing CMRT to dose-escalated EBRT have demonstrated superior biochemical control with CMRT. Longer followup of the randomized trials will be required to determine if this will translate to a benefit in metastasis-free survival, disease-specific survival, and overall survival. Although greater toxicity has been associated with CMRT compared to EBRT, recent studies suggest that technological advances that allow better definition and sparing of critical adjacent structures as well as increasing experience with brachytherapy have improved implant quality and the toxicity profile of brachytherapy. The role of androgen deprivation therapy is well established in the external beam literature for high-risk disease, but there is controversy regarding the applicability of these data in the setting of dose escalation. At this time, there is not sufficient evidence for the omission of androgen deprivation therapy with dose escalation in this population. Comparisons with surgery remain limited by differences in patient selection, but the evidence would suggest better disease control with CMRT compared to surgery alone. Due to a series of technological advances, modern combination series have demonstrated unparalleled rates of disease control in the high-risk population. Given the evidence from recent randomized trials, combination therapy may become the standard of care for high-risk cancers. Copyright © 2016 American Brachytherapy Society. Published by Elsevier Inc. All

  14. High-dose rate brachytherapy (HDRB) for primary or recurrent cancer in the vagina

    International Nuclear Information System (INIS)

    Beriwal, Sushil; Heron, Dwight E; Mogus, Robert; Edwards, Robert P; Kelley, Joseph L; Sukumvanich, Paniti

    2008-01-01

    The purpose of this study was to evaluate the efficacy of HDR brachytherapy for primary or recurrent vaginal cancer. Between the years 2000 to 2006, 18 patients with primary or recurrent vaginal cancer were treated with brachytherapy (HDRB). Six patients had primary vaginal cancer (stage II to IVA) while 12 were treated for isolated vaginal recurrence (primary cervix = 4, vulva = 1 and endometrium = 7). Five patients had previous pelvic radiation therapy. All except one patient received external beam radiation therapy to a median dose of 45 Gy (range 31.2–55.8 Gy). The HDRB was intracavitary using a vaginal cylinder in 5 patients and interstitial using a modified Syed-Nesblett template in 13 patients. The dose of interstitial brachytherapy was 18.75 Gy in 5 fractions delivered twice daily. The median follow-up was 18 months (range 6–66 months). Complete response (CR) was achieved in all but one patient (94%). Of these 17 patients achieving a CR, 1 had local recurrence and 3 had systemic recurrence at a median time of 6 months (range 6–22 months). The 2-year actuarial local control and cause-specific survival for the entire group were 88% and 82.5%, respectively. In subset analysis, the crude local control was 100% for primary vaginal cancer, 100% for the group with recurrence without any prior radiation and 67% for group with recurrence and prior radiation therapy. Two patients had late grade 3 or higher morbidity (rectovaginal fistula in one patient and chronic vaginal ulcer resulting in bleeding in one patient). Both these patients had prior radiation therapy. Our small series suggests that HDRB is efficacious for primary or recurrent vaginal cancer. Patients treated with primary disease and those with recurrent disease without prior irradiation have the greatest benefit from HDRB in this setting. The salvage rate for patients with prior radiation therapy is lower with a higher risk of significant complications. Additional patients and follow-up are ongoing

  15. Electron microscopy studies of ion implanted silicon

    International Nuclear Information System (INIS)

    Seshan, K.

    1975-11-01

    The nature of defects resulting from the implantation of phosphorous ions into doped silicon and a model of how they form are reported. This involved an electron microscope study of the crystallographic defects (in the 300A size range in concentration of 10 15 /cm 3 ) that form upon annealing. Images formed by these crystallographic defects are complex and that nonconventional imaging techniques are required for their characterization. The images of these small defects (about 300A) are sensitive to various parameters, such as foil thickness, their position in the foil, and diffracting conditions. The defects were found to be mostly interstitial hexagonal Frank loops lying on the four [111] planes and a few perfect interstitial loops; these loops occurred in concentrations of about 10 16 /cm 3 . In addition, ''rod like'' linear defects that are shown to be interstitial are also found in concentrations of 10 13 /cm 3 . It was found that the linear defects require boron for their formation. A model is proposed to account for the interstitial defects. The number of point defects that make up the defects is of the same order as the number of implanted ions. The model predicts that only interstitial loops ought to be observed in agreement with several recent investigations. Dislocation models of the loops are examined and it is shown that phosphorous ions could segregate to the Frank loops, changing their displacement vectors to a/x[111]. (x greater than 3) thus explaining the contrast effects observed. It would also explain the relative electrical inactivity of P + ion implants

  16. Argon plasma coagulation for rectal bleeding after prostate brachytherapy

    International Nuclear Information System (INIS)

    Smith, Stephen; Wallner, Kent; Dominitz, Jason A.; Han, Ben; True, Lawrence; Sutlief, Steven; Billingsley, Kevin

    2001-01-01

    Purpose: To better define the efficacy and safety of argon plasma coagulation (APC), specifically for brachytherapy-related proctitis, we reviewed the clinical course of 7 patients treated for persistent rectal bleeding. Approximately 2-10% of prostate cancer patients treated with 125 I or 103 Pd brachytherapy will develop radiation proctitis. The optimum treatment for patients with persistent bleeding is unclear from the paucity of available data. Prior reports lack specific dosimetric information, and patients with widely divergent forms of radiation were grouped together in the analyses. Methods and Materials: Seven patients were treated with APC at the Veterans Affairs Puget Sound Health Care System and the University of Washington from 1997 to 1999 for persistent rectal bleeding due to prostate brachytherapy-related proctitis. Four patients received supplemental external beam radiation, delivered by a four-field technique. A single gastroenterologist at the Veterans Affairs Puget Sound Health Care System treated 6 of the 7 patients. If the degree of proctitis was limited, all sites of active bleeding were coagulated in symptomatic patients. An argon plasma coagulator electrosurgical system was used to administer treatments every 4-8 weeks as needed. The argon gas flow was set at 1.6 L/min, with an electrical power setting of 40-45 W. Results: The rectal V100 (the total rectal volume, including the lumen, receiving the prescription dose or greater) for the 7 patients ranged from 0.13 to 4.61 cc. Rectal bleeding was first noticed 3-18 months after implantation. APC (range 1-3 sessions) was performed 9-22 months after implantation. Five patients had complete resolution of their bleeding, usually within days of completing APC. Two patients had only partial relief from bleeding, but declined additional APC therapy. No patient developed clinically evident progressive rectal wall abnormalities after APC, (post-APC follow-up range 4-13 months). Conclusions: Most

  17. Brachytherapy and percutaneous stenting in the treatment of cholangiocarcinoma: A prospective randomised study

    International Nuclear Information System (INIS)

    Valek, Vlastimil; Kysela, Petr; Kala, Zdenek; Kiss, Igor; Tomasek, Jiri; Petera, Jiri

    2007-01-01

    Purpose: To evaluate the effect of radiation therapy including intraluminal brachyterapy with iridium-192 on survival of patients with malignant biliary strictures (cholangiocarcinoma, histologically improved) treated with metallic stent in a prospective randomised study. Method and materials: In the prospective randomised study, 21 patients with cholangiocarcinoma were treated with implantation of percutaneous stents followed with intraluminal Ir-192 brachytherapy (mean dose 30 Gy) and external radiotherapy (mean dose 50 Gy) and 21 patients were treated only with stents insertion. We did not find any statistically significant differences in age and tumor localization between these two groups of patients. Results: All the patients died. In the group of patients treated with brachytherapy and with stent implantation, the mean survival time was 387.9 days. In the group of patients treated only with stent insertion the mean survival was 298 days. In effort to eliminate possible effect of external radiotherapy we treated the control group of eight patients with cholangiocarcinoma by stent insertion and brachyterapy only. Conclusion: Our results show that combined radiation therapy could extend the survival in the patients with cholangiocarcinoma obstruction

  18. Long term conservation of sexual functions after prostate brachytherapy by permanents implants

    International Nuclear Information System (INIS)

    Pena, P.C.; Hijal, T.; Pierrat, N.; Pontvert, D.; Cosset, J.M.; Thiounn, N.; Flam, T.; Chauveinc, L.

    2009-01-01

    These series suggest that the preservation of sexual abilities after prostate brachytherapy would be in relation with the previous performances and age. So, in patients aged over seventy years and with an satisfying initial IIEF5 score, the conservation rate at long term appears to go over the 50%. (N.C.)

  19. Brachytherapy Combined With Surgery for Conservative Treatment of Children With Bladder Neck and/or Prostate Rhabdomyosarcoma

    Energy Technology Data Exchange (ETDEWEB)

    Chargari, Cyrus, E-mail: cyrus.chargari@gustaveroussy.fr [Brachytherapy Unit, Department of Radiotherapy, Gustave Roussy, Villejuif (France); Institut de Recherche Biomédicale des Armées, Bretigny-sur-Orge (France); French Military Health Service Academy, Ecole du Val-de-Grâce, Paris (France); Haie-Meder, Christine [Brachytherapy Unit, Department of Radiotherapy, Gustave Roussy, Villejuif (France); Guérin, Florent [Department of Pediatric Surgery, Bicêtre Hospital, Hôpitaux Universitaires Paris Sud, Assistance Publique des Hôpitaux de Paris, Le Kremlin-Bicêtre (France); Minard-Colin, Véronique [Department of Pediatric and Adolescent Oncology, Gustave Roussy, Villejuif (France); Lambert, Guénolée de [Department of Pediatric Surgery, Bicêtre Hospital, Hôpitaux Universitaires Paris Sud, Assistance Publique des Hôpitaux de Paris, Le Kremlin-Bicêtre (France); Mazeron, Renaud; Escande, Alexandre; Marsolat, Fanny; Dumas, Isabelle [Brachytherapy Unit, Department of Radiotherapy, Gustave Roussy, Villejuif (France); Deutsch, Eric [Brachytherapy Unit, Department of Radiotherapy, Gustave Roussy, Villejuif (France); Faculté de Médecine Paris Sud, Université Paris Sud, Université Paris Saclay, Paris (France); Valteau-Couanet, Dominique [Department of Pediatric and Adolescent Oncology, Gustave Roussy, Villejuif (France); and others

    2017-06-01

    Purpose: To report the results of a conservative strategy based on partial surgery combined with brachytherapy in a prospective cohort of children with bladder–prostate rhabdomyosarcoma (BP RMS). Methods and Materials: We prospectively documented the outcome of children treated in our department between 1991 and 2015 for BP RMS and undergoing a multimodal approach combining conservative surgery (partial cystectomy and/or partial prostatectomy) and perioperative interstitial low-dose-rate or pulse-dose-rate brachytherapy. Before brachytherapy, children had received chemotherapy with modalities depending on their risk group of treatment. Results: A total of 100 patients were identified, with a median age of 28 months (range, 5.6 months-14 years). According to the Intergroup Rhabdomyosarcoma Study (IRS) group, 84 were IRS-III, and 12 were IRS-IV tumors. Four patients were treated at relapse. The median number of chemotherapy cycles before local therapy was 6 (range, 4-13). After surgery, 63 patients had a macroscopic tumor residuum. Five patients underwent a brachytherapy boost before pelvic external beam radiation therapy because of nodal involvement, and 95 had exclusive brachytherapy. Median follow-up was 64 months (range, 6 months-24.5 years). Five-year disease-free and overall survival rates were 84% (95% confidence interval 80%-88%) and 91% (95% confidence interval 87%-95%), respectively. At last follow-up most survivors presented with only mild to moderate genitourinary sequelae and a normal diurnal urinary continence. Five patients required a secondary total cystectomy: 3 for a nonfunctional bladder and 2 for relapse. Conclusion: Brachytherapy is effective as part of a conservative strategy for BP RMS, with a relatively low delayed toxicity as compared with previously published studies using external beam radiation therapy. Longer follow-up is required to ensure that the functional results are maintained over time.

  20. Brachytherapy Combined With Surgery for Conservative Treatment of Children With Bladder Neck and/or Prostate Rhabdomyosarcoma

    International Nuclear Information System (INIS)

    Chargari, Cyrus; Haie-Meder, Christine; Guérin, Florent; Minard-Colin, Véronique; Lambert, Guénolée de; Mazeron, Renaud; Escande, Alexandre; Marsolat, Fanny; Dumas, Isabelle; Deutsch, Eric; Valteau-Couanet, Dominique

    2017-01-01

    Purpose: To report the results of a conservative strategy based on partial surgery combined with brachytherapy in a prospective cohort of children with bladder–prostate rhabdomyosarcoma (BP RMS). Methods and Materials: We prospectively documented the outcome of children treated in our department between 1991 and 2015 for BP RMS and undergoing a multimodal approach combining conservative surgery (partial cystectomy and/or partial prostatectomy) and perioperative interstitial low-dose-rate or pulse-dose-rate brachytherapy. Before brachytherapy, children had received chemotherapy with modalities depending on their risk group of treatment. Results: A total of 100 patients were identified, with a median age of 28 months (range, 5.6 months-14 years). According to the Intergroup Rhabdomyosarcoma Study (IRS) group, 84 were IRS-III, and 12 were IRS-IV tumors. Four patients were treated at relapse. The median number of chemotherapy cycles before local therapy was 6 (range, 4-13). After surgery, 63 patients had a macroscopic tumor residuum. Five patients underwent a brachytherapy boost before pelvic external beam radiation therapy because of nodal involvement, and 95 had exclusive brachytherapy. Median follow-up was 64 months (range, 6 months-24.5 years). Five-year disease-free and overall survival rates were 84% (95% confidence interval 80%-88%) and 91% (95% confidence interval 87%-95%), respectively. At last follow-up most survivors presented with only mild to moderate genitourinary sequelae and a normal diurnal urinary continence. Five patients required a secondary total cystectomy: 3 for a nonfunctional bladder and 2 for relapse. Conclusion: Brachytherapy is effective as part of a conservative strategy for BP RMS, with a relatively low delayed toxicity as compared with previously published studies using external beam radiation therapy. Longer follow-up is required to ensure that the functional results are maintained over time.

  1. Adjuvant iodine-125 brachytherapy for hepatocellular carcinoma after complete hepatectomy: a randomized controlled trial.

    Directory of Open Access Journals (Sweden)

    Kaiyun Chen

    Full Text Available BACKGROUND: Tumor recurrence is a major problem after curative resection of hepatocellular carcinoma (HCC. The current study evaluated the effects of adjuvant iodine-125 ((125I brachytherapy on postoperative recurrence of HCC. METHODOLOGY/PRINCIPAL FINDINGS: From July 2000 to June 2004, 68 HCC patients undergoing curative hepatectomy were randomly assigned into a (125I adjuvant brachytherapy group (n = 34 and a group of best care (n = 34. Patients in the (125I adjuvant brachytherapy group received (125I seed implantation on the raw surface of resection. Patients in the best care control group received identical treatments except for the (125I seed implantation. Time to recurrence (TTR and 1-, 3- and 5-year overall survival (OS were compared between the two groups. The follow-up ended in January 2010, and lasted for 7.7-106.4 months with a median of 47.6 months. TTR was significantly longer in the (125I group (mean of 60.0 months vs. 36.7 months in the control. The 1-, 3- and 5-year recurrence-free rates of the (125I group were 94.12%, 76.42%, and 73.65% vs. 88.24%, 50.00%, and 29.41% compared with the control group, respectively. The 1-, 3- and 5-year OS rates of the (125I group were 94.12%, 73.53%, and 55.88% vs. 88.24%, 52.94%, and 29.41% compared with the control group, respectively. The (125I brachytherapy decreased the risk of recurrence (HR = 0.310 and the risk of death (HR = 0.364. Most frequent adverse events in the (125I group included nausea, vomiting, arrhythmia, decreased white blood cell and/or platelet counts, and were generally mild and manageable. CONCLUSIONS/SIGNIFICANCE: Adjuvant (125I brachytherapy significantly prolonged TTR and increased the OS rate after curative resection of HCC. TRIAL REGISTRATION: Australian New Zealand Clinical Trials Registry ACTRN12610000081011.

  2. SU-F-BRA-04: Prostate HDR Brachytherapy with Multichannel Robotic System

    International Nuclear Information System (INIS)

    Joseph, F Maria; Podder, T; Yu, Y

    2015-01-01

    Purpose: High-dose-rate (HDR) brachytherapy is gradually becoming popular in treating patients with prostate cancers. However, placement of the HDR needles at desired locations into the patient is challenging. Application of robotic system may improve the accuracy of the clinical procedure. This experimental study is to evaluate the feasibility of using a multichannel robotic system for prostate HDR brachytherapy. Methods: In this experimental study, the robotic system employed was a 6-DOF Multichannel Image-guided Robotic Assistant for Brachytherapy (MIRAB), which was designed and fabricated for prostate seed implantation. The MIRAB has the provision of rotating 16 needles while inserting them. Ten prostate HDR brachytherapy needles were simultaneously inserted using MIRAB into a commercially available prostate phantom. After inserting the needles into the prostate phantom at desired locations, 2mm thick CT slices were obtained for dosimetric planning. HDR plan was generated using Oncetra planning system with a total prescription dose of 34Gy in 4 fractions. Plan quality was evaluated considering dose coverage to prostate and planning target volume (PTV), with 3mm margin around prostate, as well as the dose limit to the organs at risk (OARs) following the American Brachytherapy Society (ABS) guidelines. Results: From the CT scan, it is observed that the needles were inserted straight into the desired locations and they were adequately spaced and distributed for a clinically acceptable HDR plan. Coverage to PTV and prostate were about 91% (V100= 91%) and 96% (V100=96%), respectively. Dose to 1cc of urethra, rectum, and bladder were within the ABS specified limits. Conclusion: The MIRAB was able to insert multiple needles simultaneously into the prostate precisely. By controlling the MIRAB to insert all the ten utilized needles into the prostate phantom, we could achieve the robotic HDR brachytherapy successfully. Further study for assessing the system

  3. Use of a Flexible Inflatable Multi-Channel Applicator for Vaginal Brachytherapy in the Management of Gynecologic Cancer

    Directory of Open Access Journals (Sweden)

    Samuel M Shin

    2015-09-01

    Full Text Available Introduction: Evaluate use of novel multi-channel applicator (MC CapriTM to improve vaginal disease coverage achievable by single-channel applicator (SC and comparable to Syed plan simulation. Material and Methods: 28 plans were evaluated from 4 patients with primary or recurrent gynecologic cancer in the vagina. Each received whole pelvis radiation, followed by 3 weekly treatments using HDR brachytherapy with a 13-channel MC. Upper vagina was treated to 5 mm depth to 1500 cGy/3 fractions with a simultaneous integrated boost totaling 2100 cGy/3 fractions to tumor. Modeling of SC and Syed plans was performed using MC scans for each patient. Dosimetry for MC and SC plans was evaluated for PTV700 cGy coverage, maximum dose to 2cm3 to bladder, rectum as well as mucosal surface points. Dosimetry for Syed plans was calculated for PTV700 cGy coverage. Patients were followed for treatment response and toxicity.Results: Dosimetric analysis between MC and SC plans demonstrated increased tumor coverage (PTV700 cGy, with decreased rectal, bladder, and contralateral vaginal mucosa dose in favor of MC. These differences were significant (p<0.05. Comparison of MC and Syed plans demonstrated increased tumor coverage in favor of Syed plans which were not significant (p=0.71. Patients treated with MC had no cancer recurrence or ≥ grade 3 toxicity.Conclusion: Use of MC was efficacious and safe, providing superior coverage of tumor volumes ≤1cm depth compared to SC and comparable to Syed implant. MC avoids excess dose to surrounding organs compared to SC, and potentially less morbidity than Syed implants. For tumors extending ≤1cm depth, use of MC represents an alternative to an interstitial implant.

  4. American brachytherapy society (ABS) consensus guidelines for brachytherapy of esophageal cancer

    International Nuclear Information System (INIS)

    Gaspar, Laurie E.; Nag, Subir; Herskovic, Arnold; Mantravadi, Rao; Speiser, Burton

    1997-01-01

    Introduction: There is wide variation in the indications, treatment regimens, and dosimetry for brachytherapy in the treatment of cancer of the esophagus. No guidelines for optimal therapy currently exist. Methods and Materials: Utilizing published reports and clinical experience, representatives of the Clinical Research Committee of the American Brachytherapy Society (ABS) formulated guidelines for brachytherapy in esophageal cancer. Results: Recommendations were made for brachytherapy in the definitive and palliative treatment of esophageal cancer. (A) Definitive treatment: Good candidates for brachytherapy include patients with unifocal thoracic adeno- or squamous cancers ≤ 10 cm in length, with no evidence of intra-abdominal or metastatic disease. Contraindications include tracheal or bronchial involvement, cervical esophagus location, or stenosis that cannot be bypassed. The esophageal brachytherapy applicator should have an external diameter of 6-10 mm. If 5FU-based chemotherapy and 45-50-Gy external beam are used, recommended brachytherapy is either: (i) HDR 10 Gy in two weekly fractions of 5 Gy each; or (ii) LDR 20 Gy in a single course at 0.4-1 Gy/hr. All doses are specified 1 cm from the midsource or middwell position. Brachytherapy should follow external beam radiation therapy and should not be given concurrently with chemotherapy. (B) Palliative treatment: Patients with adeno- or squamous cancers of the thoracic esophagus with distant metastases or unresectable local disease progression/recurrence after definitive radiation treatment should be considered for brachytherapy with palliative intent. After limited dose (30 Gy) EBRT, the recommended brachytherapy is either: (i) HDR 10-14 Gy in one or two fractions; or (ii) LDR 20-25 Gy in a single course at 0.4-1 Gy/hr. The need for external beam radiation in newly diagnosed patients with a life expectancy of less than 3 months is controversial. In these cases, HDR of 15-20 Gy in two to four fractions or

  5. Brachytherapy dose-volume histogram computations using optimized stratified sampling methods

    International Nuclear Information System (INIS)

    Karouzakis, K.; Lahanas, M.; Milickovic, N.; Giannouli, S.; Baltas, D.; Zamboglou, N.

    2002-01-01

    A stratified sampling method for the efficient repeated computation of dose-volume histograms (DVHs) in brachytherapy is presented as used for anatomy based brachytherapy optimization methods. The aim of the method is to reduce the number of sampling points required for the calculation of DVHs for the body and the PTV. From the DVHs are derived the quantities such as Conformity Index COIN and COIN integrals. This is achieved by using partial uniform distributed sampling points with a density in each region obtained from a survey of the gradients or the variance of the dose distribution in these regions. The shape of the sampling regions is adapted to the patient anatomy and the shape and size of the implant. For the application of this method a single preprocessing step is necessary which requires only a few seconds. Ten clinical implants were used to study the appropriate number of sampling points, given a required accuracy for quantities such as cumulative DVHs, COIN indices and COIN integrals. We found that DVHs of very large tissue volumes surrounding the PTV, and also COIN distributions, can be obtained using a factor of 5-10 times smaller the number of sampling points in comparison with uniform distributed points

  6. Interstitial brachytherapy for stage I and II squamous cell carcinoma of the oral tongue: factors influencing local control and soft tissue complications

    International Nuclear Information System (INIS)

    Fujita, Minoru; Hirokawa, Yutaka; Kashiwado, Kouzo; Akagi, Yukio; Kashimoto, Kazuki; Kiriu, Hiroshi; Matsuura, Kanji; Ito, Katsuhide

    1999-01-01

    Purpose: Our aim was to study the treatment parameters that influence local control and soft tissue complications (STC) in a series of 207 Stage I and II squamous cell carcinomas of the oral tongue treated by interstitial brachytherapy (BRT) alone (127 patients), or by a combination using external beam irradiation (EBI) (80 patients) between 1980 and 1993. Methods and Materials: The patient distribution was 93 T1, 72 T2a, and 42 T2b. The prescribed BRT dose at the plane 5 mm from the plane of the radioactive sources was 65-70 Gy in BRT alone, and 50-60 Gy in the combined treatment using EBI. Generally, an EBI dose of 30 Gy was used. No prophylactic neck treatment was performed. Results: The 5-year local recurrence-free rate for T1, T2a, and T2b was 92.9%, 81.9%, and 71.8%, respectively (p 85 Gy (p = 0.01), BRT dose > 55 Gy (p = 0.04), and a dose rate = 0.6 Gy/h was significantly related to more STC in BRT alone (p = 0.03), and that a dose rate > = 0.55 Gy/h (p 70 Gy (p 100 Gy (p 85 Gy and 55 Gy and < = 70 Gy, and < 0.55 Gy/h, respectively. We also recommend longer follow-up periods; more than 5 years might be necessary for late local recurrences and for secondary cancers

  7. Novel Use of the Contura for High Dose Rate Cranial Brachytherapy

    International Nuclear Information System (INIS)

    Scanderbeg, Daniel J.; Alksne, John F.; Lawson, Joshua D.; Murphy, Kevin T.

    2011-01-01

    A popular choice for treatment of recurrent gliomas was cranial brachytherapy using the GliaSite Radiation Therapy System. However, this device was taken off the market in late 2008, thus leaving a treatment void. This case study presents our experience treating a cranial lesion for the first time using a Contura multilumen, high-dose-rate (HDR) brachytherapy balloon applicator. The patient was a 47-year-old male who was diagnosed with a recurrent right frontal anaplastic oligodendroglioma. Previous radiosurgery made him a good candidate for brachytherapy. An intracavitary HDR balloon brachytherapy device (Contura) was placed in the resection cavity and treated with a single fraction of 20 Gy. The implant, treatment, and removal of the device were all completed without incident. Dosimetry of the device was excellent because the dose conformed very well to the target. V90, V100, V150, and V200 were 98.9%, 95.7%, 27.2, and 8.8 cc, respectively. This patient was treated successfully using the Contura multilumen balloon. Contura was originally designed for deployment in a postlumpectomy breast for treatment by accelerated partial breast irradiation. Being an intracavitary balloon device, its similarity to the GliaSite system makes it a viable replacement candidate. Multiple lumens in the device also make it possible to shape the dose delivered to the target, something not possible before with the GliaSite applicator.

  8. Radiological response of ceramic and polymeric devices for breast brachytherapy

    International Nuclear Information System (INIS)

    Batista Nogueira, Luciana; Passos Ribeiro de Campos, Tarcisio

    2012-01-01

    In the present study, the radiological visibility of ceramic and polymeric devices implanted in breast phantom was investigated for future applications in brachytherapy. The main goal was to determine the radiological viability of ceramic and polymeric devices in vitro by performing simple radiological diagnostic methods such as conventional X-ray analysis and mammography due to its easy access to the population. The radiological response of ceramic and polymeric devices implanted in breast phantom was determined using conventional X-ray, mammography and CT analysis. - Highlights: ► Radiological visibility of ceramic and polymeric devices implanted in breast phantom. ► The barium incorporation in the seed improves the radiological contrast. ► Radiological monitoring shows the position, orientation and degradation of devices. ► Simple radiological methods such as X-ray and mammography were used for radiological monitoring.

  9. Intracavitary mould brachytherapy in malignant tumors of the maxilla

    International Nuclear Information System (INIS)

    Rosenblatt, Edward; Blumenfeld, Israel; Cederbaum, Martin; Kuten, Abraham

    1996-01-01

    Purpose: To integrate brachytherapy in the combined modality management of malignant tumors of the maxilla, as a means of increasing the radiotherapy dose to the tumor bed while avoiding high doses to the orbital contents. Materials and methods: Following a partial or total maxillectomy, a duplication of the interim surgical obturator was created using a wash of vinyl polysiloxane. This mould was used as a carrier for afterloading nylon catheters through which 192-Iridium seed-ribbons were inserted. Following brachytherapy, selected patients also received external beam irradiation. Results and discussion: After a median follow-up of 36 months, 9 out of 11 patients are alive and disease-free; 1 developed a local recurrence and another relapsed at another site in the oral cavity. Transient grade 1 - 2 mucositis at the implant site was observed in all patients. The review of computer isodose distributions showed that the average dose received by the homolateral eyeball was 10% (range 9,2 - 10.0) of the prescribed surface dose to the surgical cavity. Conclusions: Brachytherapy can be integrated in the management of patients with malignant tumors of the maxilla in the form of a custom-made intracavitary mould carrying 192-Iridium sources. We found this technique particularly useful in cases with close or positive surgical margins

  10. American Society for Radiation Oncology (ASTRO) and American College of Radiology (ACR) Practice Guideline for the Performance of High-Dose-Rate Brachytherapy

    International Nuclear Information System (INIS)

    Erickson, Beth A.; Demanes, D. Jeffrey; Ibbott, Geoffrey S.; Hayes, John K.; Hsu, I-Chow J.; Morris, David E.; Rabinovitch, Rachel A.; Tward, Jonathan D.; Rosenthal, Seth A.

    2011-01-01

    High-Dose-Rate (HDR) brachytherapy is a safe and efficacious treatment option for patients with a variety of different malignancies. Careful adherence to established standards has been shown to improve the likelihood of procedural success and reduce the incidence of treatment-related morbidity. A collaborative effort of the American College of Radiology (ACR) and American Society for Therapeutic Radiation Oncology (ASTRO) has produced a practice guideline for HDR brachytherapy. The guideline defines the qualifications and responsibilities of all the involved personnel, including the radiation oncologist, physicist and dosimetrists. Review of the leading indications for HDR brachytherapy in the management of gynecologic, thoracic, gastrointestinal, breast, urologic, head and neck, and soft tissue tumors is presented. Logistics with respect to the brachytherapy implant procedures and attention to radiation safety procedures and documentation are presented. Adherence to these practice guidelines can be part of ensuring quality and safety in a successful HDR brachytherapy program.

  11. Radiation safety parameters following prostate brachytherapy

    International Nuclear Information System (INIS)

    Smathers, Sesalie; Wallner, Kent; Korssjoen, Tammy; Bergsagel, Carl; Hudson, Rick H.; Sutlief, Steven; Blasko, John

    1999-01-01

    Purpose: To determine the degree and variability of radiation exposure to the general public from patients after I-125 or Pd-103 prostate brachytherapy. Methods and Materials: Radiation exposure measurements were made from 38 consecutive, unselected patients with stage T1 or T2 prostatic carcinoma who had trans perineal I-125 or Pd-103 implants at the University of Washington in 1998. Results: The exposure rate at the anterior skin surface following a I-125 implant ranged from 2.2 to 8.9 mrem/hour (average: 5.0). The exposure rate at the anterior skin surface from a Pd-103 implant ranged from 0.5 to 4.9 mrem/hour (average: 1.7). Based on the current Nuclear Regulatory Commission (NRC) regulations the time required to reach the annual limit at the anterior skin surface would be 20 hours for I-125 and 59 hours for Pd-103. For exposure at the lateral skin surface, the times would exceed 500 hours for either isotope. Conclusions: This data suggest that patients need not be concerned about being a radiation risk to the general public following their procedure

  12. First symposium seed implant 125I and high rate of prostate

    International Nuclear Information System (INIS)

    2012-01-01

    The First symposium seed implant 125 I and high rate of prostate, was organized by the Marie Curie Foundation, between the 12 to april 2012, in the Cordoba city of Argentina. In this event were presented several documents in different topics: patients selection for impacts of 125 I seeds; high doses radiation in radiotherapy; brachytherapy for prostate cancer; prostate implant technique with 125 I seeds; implant dosimetric aspects; radioprotection of 125 I seeds.

  13. Results of brachytherapy boost in high risk breast cancer patients

    International Nuclear Information System (INIS)

    Battermann, J.J.

    1996-01-01

    Introduction: in breast conserving therapy the role of brachytherapy as a boost after whole breast irradiation is not clear. The series from the Netherlands Cancer Institute show a very high local control rate, but the question could be raised whether all these patients need a brachy boost. Therefore, it was decided at our institute, to deliver a brachy boost only to high risk patients, viz. patients with incomplete resection margins and/or extensive in situ cancer (ECI). Materials and methods: in the period 1988 through 1993 a total of 148 patients with 151 breast tumours received a boost on the tumour bed using brachytherapy. Age varied from 25 till 74 years, with a mean age of 52.3 years. Incomplete resection margins were found in 60 patients, ECI in 31 and both in 49 patients. In the majority of patients, the ECI component was not completely removed. T-stage was unknown in 9 patients. T1 in 83, T2 in 49 and T3 in 10. Nodal status was N0 in 119 and N1 in 33 patients. Infiltrating duct carcinoma was the most common histology. No infiltrating growth was found in 6 patients, but one patient presented a positive node. The interval period between day of operation and day of brachytherapy implantation was between 3 and 4 months in 62%. The mean interval between completion of beam irradiation and day of implantation was 18 days, while 12 patients received their brachytherapy previous to the beam irradiation. External irradiation was with two tangential fields and a total dose of 50 Gy in 25 fractions over 6 weeks (9 fractions in two weeks). The number of needles in two planes. Most patients were implanted under local anaesthesia. Dose rate in 97 patients was 51 - 60 cGy/h. Results: follow-up for patients alive varied from 2 years till 7 years with a mean follow-up period of 4 years. One hundred and twenty five patients are alive, including 6 patients with manifest metastases. Local recurrence was encountered in 8 patients (interval 14 - 60 months, mean 30 months), with

  14. MO-D-BRD-03: Radiobiology and Commissioning of Electronic Brachytherapy for IORT

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, J. [Oregon Health & Science Univ (United States)

    2015-06-15

    Electronic brachytherapy (eBT) has seen an insurgence of manufacturers entering the US market for use in radiation therapy. In addition to the established interstitial, intraluminary, and intracavitary applications of eBT, many centers are now using eBT to treat skin lesions. It is important for medical physicists working with electronic brachytherapy sources to understand the basic physics principles of the sources themselves as well as the variety of applications for which they are being used. The calibration of the sources is different from vendor to vendor and the traceability of calibrations has evolved as new sources came to market. In 2014, a new air-kerma based standard was introduced by the National Institute of Standards and Technology (NIST) to measure the output of an eBT source. Eventually commercial treatment planning systems should accommodate this new standard and provide NIST traceability to the end user. The calibration and commissioning of an eBT system is unique to its application and typically entails a list of procedural recommendations by the manufacturer. Commissioning measurements are performed using a variety of methods, some of which are modifications of existing AAPM Task Group protocols. A medical physicist should be familiar with the different AAPM Task Group recommendations for applicability to eBT and how to properly adapt them to their needs. In addition to the physical characteristics of an eBT source, the photon energy is substantially lower than from HDR Ir-192 sources. Consequently, tissue-specific dosimetry and radiobiological considerations are necessary when comparing these brachytherapy modalities and when making clinical decisions as a radiation therapy team. In this session, the physical characteristics and calibration methodologies of eBt sources will be presented as well as radiobiology considerations and other important clinical considerations. Learning Objectives: To understand the basic principles of electronic

  15. WE-DE-201-08: Multi-Source Rotating Shield Brachytherapy Apparatus for Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Dadkhah, H; Wu, X [University of Iowa, Iowa City, Iowa (United States); Kim, Y; Flynn, R [University of Iowa Hospitals and Clinics, Iowa City, IA (United States)

    2016-06-15

    Purpose: To introduce a novel multi-source rotating shield brachytherapy (RSBT) apparatus for the precise simultaneous angular and linear positioning of all partially-shielded 153Gd radiation sources in interstitial needles for treating prostate cancer. The mechanism is designed to lower the detrimental dose to healthy tissues, the urethra in particular, relative to conventional high-dose-rate brachytherapy (HDR-BT) techniques. Methods: Following needle implantation, the delivery system is docked to the patient template. Each needle is coupled to a multi-source afterloader catheter by a connector passing through a shaft. The shafts are rotated by translating a moving template between two stationary templates. Shaft walls as well as moving template holes are threaded such that the resistive friction produced between the two parts exerts enough force on the shafts to bring about the rotation. Rotation of the shaft is then transmitted to the shielded source via several keys. Thus, shaft angular position is fully correlated with the position of the moving template. The catheter angles are simultaneously incremented throughout treatment as needed, and only a single 360° rotation of all catheters is needed for a full treatment. For each rotation angle, source depth in each needle is controlled by a multi-source afterloader, which is proposed as an array of belt-driven linear actuators, each of which drives a source wire. Results: Optimized treatment plans based on Monte Carlo dose calculations demonstrated RSBT with the proposed apparatus reduced urethral D{sub 1cc} below that of conventional HDR-BT by 35% for urethral dose gradient volume within 3 mm of the urethra surface. Treatment time to deliver 20 Gy with multi-source RSBT apparatus using nineteen 62.4 GBq {sup 153}Gd sources is 117 min. Conclusions: The proposed RSBT delivery apparatus in conjunction with multiple nitinol catheter-mounted platinum-shielded {sup 153}Gd sources enables a mechanically feasible

  16. A phase I/II clinical trial for the hybrid of intracavitary and interstitial brachytherapy for locally advanced cervical cancer.

    Science.gov (United States)

    Murakami, Naoya; Kato, Shingo; Nakano, Takashi; Uno, Takashi; Yamanaka, Takeharu; Sakurai, Hideyuki; Yoshimura, Ryoichi; Hiratsuka, Junichi; Kuroda, Yuki; Yoshio, Kotaro; Itami, Jun

    2016-08-17

    This paper describes about a study protocol of phase I/II multicenter prospective clinical trial evaluating the feasibility and efficacy of the hybrid of intracavitary and interstitial brachytherapy (HBT) for locally advanced uterine cervical cancer patients. Patients with histologically confirmed FIGO stage IB2, IIA2, IIB, and IIIB uterine cervical carcinoma width of which is larger than 5 cm assessed by MRI will be entered to this clinical trial. Protocol therapy is 30-30.6 Gy in 15-17 fractions of whole pelvic radiotherapy concurrent with weekly CDDP (40 mg/m(2)), followed by 24 Gy in 4 fractions of HBT and central shield EBRT up to 50-50.4 Gy in 25-28 fractions. Tumor width is assessed again within one week before the first HBT and if the tumor width is larger than 4 cm, patients proceed to the secondary registration. In phase I section, feasibility of this will be investigated. If less than 10 % out of 20 patients experienced greater than grade 3 acute non-hematologic adverse effects, the study proceeds to phase II part. In phase II part a total of 55 patients will be accrued and the efficacy of the HBT will be investigated comparing with historical control data. If the lower margin of 90 % confidence interval of the 2-year pelvic progression-free survival of the HBT trial is higher than 64 %, the HBT is considered to be more effective than conventional ICBT. The aim of this study is to demonstrate the feasibility and efficacy of the HBT for locally advanced cervical cancer. This trial will clarify the indication, feasibility, and efficacy of this new technique. UMIN000019081 ; Registration date: 2015/9/30.

  17. A Japanese prospective multi-institutional feasibility study on accelerated partial breast irradiation using interstitial brachytherapy: treatment planning and quality assurance

    International Nuclear Information System (INIS)

    Otani, Yuki; Nose, Takayuki; Dokiya, Takushi; Saeki, Toshiaki; Kumazaki, Yu

    2015-01-01

    In Japan, breast-conserving surgery with closed cavity has generally been performed for breast cancer patients, and accelerated partial breast irradiation (APBI) is considered difficult because Asian females generally have smaller breast sizes than Western females. Therefore, common identification of target and treatment plan method in APBI is required. A prospective multicenter study was conducted in Japan to determine institutional compliance with APBI using high-dose-rate interstitial brachytherapy (ISBT) designed for Japanese female patients. For this study, 46 patients were recruited at eight institutions from January 2009 to December 2011. The reproducibility of the ISBT–APBI plan was evaluated using three criteria: (1) minimum clinical target volume dose with a clip dose ≥ 6 Gy/fraction, (2) irradiated volume constraint of 40-150 cm 3 , and (3) uniformity of dose distribution, expressed as the dose non-uniformity ratio (DNR, V150/V100) < 0.35. The ISBT–APBI plan for each patient was considered reproducible when all three criteria were met. When the number of non-reproducible patients was ≤ 4 at study completion, APBI at this institution was considered statistically reproducible. Half of the patients (52 %) had a small bra size (A/B cup). The mean values of the dose-constrained parameters were as follows: Vref, 117 cm 3 (range, 40-282), DNR, 0.30 (range, 0.22-0.51), and clip dose, 784 cGy (range, 469-3146). A total of 43/46 treatment plans were judged to be compliant and ISBT–APBI was concluded to be reproducible. This study showed that multi-institutional ISBT–APBI treatment plan was reproducible for small breast patient with closed cavity

  18. Multicentre treatment planning study of MRI-guided brachytherapy for cervical cancer: Comparison between tandem-ovoid applicator users

    International Nuclear Information System (INIS)

    Nomden, Christel N.; Leeuw, Astrid A.C. de; Van Limbergen, Erik; Brabandere, Marisol de; Nulens, An; Nout, Remi A.; Laman, Mirjam; Ketelaars, Martijn; Lutgens, Ludovicus; Reniers, Brigitte; Jürgenliemk-Schulz, Ina Maria

    2013-01-01

    Background and purpose: To compare MRI-guided treatment planning approaches between four centres that use tandem-ovoid applicators. Material and methods: Four centres generated three treatment plans for four patients: standard, optimised intracavitary, and optimised intracavitary/interstitial. Prescribed D90 High-Risk CTV (HR-CTV) was 85 Gy EQD2 (external-beam radiotherapy and brachytherapy), while the D 2cc OAR limit was 90 Gy EQD2 for bladder and 75 Gy EQD2 for rectum, sigmoid, and bowel, respectively. DVH-parameters, source loading patterns and spatial dose distributions of the three treatment plans were compared. Results: The standard plans of the different centres were comparable with respect to the D90 HR-CTV, but differed in OAR doses. MRI-guided intracavitary optimisation resulted in organ sparing and smaller variation in DVH parameters between the centres. Adding interstitial needles led to target dose escalation while respecting the OAR constraints. However, substantial differences in relative weights of the applicator parts resulted in an increased variation in DVH parameters and locations of high dose regions. Conclusions: MRI-guided brachytherapy treatment planning optimisation provides the possibility to increase the dose to the HR-CTV and spare the OARs. Depending on the degree of conformity the centres make different choices in relative weighting of applicator parts, leading to different dose distributions

  19. Metallic artifact mitigation and organ-constrained tissue assignment for Monte Carlo calculations of permanent implant lung brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Sutherland, J. G. H.; Miksys, N.; Thomson, R. M., E-mail: rthomson@physics.carleton.ca [Carleton Laboratory for Radiotherapy Physics, Department of Physics, Carleton University, Ottawa, Ontario K1S 5B6 (Canada); Furutani, K. M. [Department of Radiation Oncology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905 (United States)

    2014-01-15

    assignment within lung contours are employed in generated phantoms, this erroneous assignment is reduced, generally resulting in higher doses. Lung-constrained tissue assignment also results in increased doses in regions of interest due to a reduction in the erroneous assignment of adipose to voxels within lung contours. Differences in dose metrics calculated for different computational phantoms are sensitive to radionuclide photon spectra with the largest differences for{sup 103}Pd seeds and smallest but still considerable differences for {sup 131}Cs seeds. Conclusions: Despite producing differences in CT images, dose metrics calculated using the STR, fan beam + STR, and 3D median filter techniques produce similar dose metrics. Results suggest that the accuracy of dose distributions for permanent implant lung brachytherapy is improved by applying lung-constrained tissue assignment schemes to metallic artifact corrected images.

  20. American brachytherapy society (ABS) guidelines for brachytherapy of esophageal cancer

    International Nuclear Information System (INIS)

    Nag, Subir; Gaspar, Laurie; Herskovic, Arnold; Mantravadi, Prasad; Speiser, Burton

    1996-01-01

    Introduction: There is wide variation in the indications, techniques, treatment regimens and dosimetry being used to treat cancer of the esophagus and no guidelines exist for optimal therapy. Methods: The Clinical Research Committee of the ABS met to formulate consensus guidelines for brachytherapy in esophageal cancer. Results: Good candidates for brachytherapy include patients with unifocal disease, with thoracic tumor 10 cm primary regional lymph adenopathy or tumor located in the gastro-esophageal junction or cervical esophagus. Contraindications include tracheo-esophageal fistula or stenosis that cannot be by-passed. The esophageal or nasogastric tube inserted should have a diameter of 6-10 mm whenever possible. If 5FU-based chemotherapy and 50 Gy external beam (EBRT) are used, it is suggested that the low dose rate brachytherapy (LDR) dose be 20 Gy at 0.4-1 Gy/hr, prescribed at 1 cm from the source. If high dose rate (HDR) is used, the dose recommended is 10 Gy in 2 weekly fractions of 5 Gy each, given after EBRT. Chemotherapy is not usually given concurrently with brachytherapy, and when it is, the brachytherapy dose is reduced. The length of esophagus treated by brachytherapy includes the post-EBRT involved area and a 1-2 cm margin proximally and distally. Supportive care, given during EBRT includes an antifungal agent (e.g., diflucan) and carafate. Gradual dilatation of the esophagus is required post-treatment for esophageal strictures. Conclusion: Guidelines were developed for brachytherapy in esophageal cancer. As more clinical data becomes available, these guidelines will be updated by the ABS

  1. Prostate-specific antigen density is predictive of outcome in suboptimal prostate seed brachytherapy.

    Science.gov (United States)

    Benzaquen, David; Delouya, Guila; Ménard, Cynthia; Barkati, Maroie; Taussky, Daniel

    In prostate seed brachytherapy, a D 90 of prostate-specific antigen + 2). Univariate and multivariate analyses were performed, adjusting for known prognostic factors such as D 90 and prostate-specific antigen density (PSAD) of ≥0.15 ng/mL/cm 3 , to evaluate their ability to predict BF. Median followup for patients without BF was 72 months (interquartile range 56-96). BF-free recurrence rate at 5 years was 95% and at 8 years 88%. In univariate analysis, PSAD and cancer of the prostate risk assessment score were predictive of BF. On multivariate analysis, none of the factors remained significant. The best prognosis had patients with a low PSAD (<0.15 ng/mL/cm 3 ) and an optimal implant at 30 days after implantation (as defined by D 90  ≥ 130 Gy) compared to patients with both factors unfavorable (p = 0.006). A favorable PSAD was associate with a good prognosis, independently of the D 90 (<130 Gy vs. ≥130 Gy, p = 0.7). Patients with a PSAD of <0.15 ng/mL/cm 3 have little risk of BF, even in the case of a suboptimal implant. These results need to be validated in other patients' cohorts. Copyright © 2016 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.

  2. Radiological protection of patients in brachytherapy

    International Nuclear Information System (INIS)

    Sacc, Ricardo; Herrero, Flavia

    2008-01-01

    Full text: The prefix 'brachy' means short-range, so brachytherapy is the administration of radiation therapy using small radioactive sources in the form of needles, tubes, wires or seeds, which are placed within the tumor -interstitial form- or very near of it, superficially or in an endo-cavity form. This technique, which was limited by the size of the primary tumor, has the advantage, that the radiation, can be adjusted to the size and shape of the tumor volume and the radioisotope used, - short range -, is selected with the criteria of getting the dose in the organs at risk, as low as possible, making what it is known as conformal radiotherapy. Radioactive sources may be permanent or temporary implants. The application of radioactive material, can be manually or automatically. In the first case, a major breakthrough from the radioprotection point of view, was the use of afterloading devices, methodology highly recommended to reduce the radiation exposure to staff. With the development of technology, remotely controlled afterloading devices were introduced, which in addition to complying with the above requirement, allow the source to move in different positions along catheters housed in one or more channels, making therapeutic brachytherapy treatments in tumor volumes possible, that due to its length, decades ago would have been an unthinkable deal. In all cases, sources, which may vary from the 3 mm in length, 125 Iodine or 198 Gold seeds, to extensive wires of 192 Iridium, are encapsulated for two main purposes: preventing leakage of radioactive material and absorption of unwanted radiation, alpha and beta, produced by the radioactive decay. Consequently, it should be highly unlikely that the radioactive material, could be lost or located in the patient, in a different place of the one that was planned. However, history shows us the opposite. Its is known the kind of deterministic effect that radiation is going to produce in the tumor, where the severity of

  3. Brachytherapy for early oral tongue cancer. Low dose rate to high dose rate

    International Nuclear Information System (INIS)

    Yamazaki, Hideya; Inoue, Takehiro; Yoshida, Ken; Yoshioka, Yasuo; Shimizutani, Kimishige; Inoue, Toshihiko; Furukawa, Souhei; Kakimoto, Naoya

    2003-01-01

    To examine the compatibility of low dose rate (LDR) with high dose rate (HDR) brachytherapy, we reviewed 399 patients with early oral tongue cancer (T1-2N0M0) treated solely by brachytherapy at Osaka University Hospital between 1967 and 1999. For patients in the LDR group (n=341), the treatment sources consisted of Ir-192 pin for 227 patients (1973-1996; irradiated dose, 61-85 Gy; median, 70 Gy), Ra-226 needle for 113 patients (1967-1986; 55-93 Gy; median, 70 Gy). Ra-226 and Ir-192 were combined for one patient. Ir-192 HDR (microSelectron-HDR) was used for 58 patients in the HDR group (1991-present; 48-60 Gy; median, 60 Gy). LDR implantations were performed via oral and HDR via a submental/submandibular approach. The dose rates at the reference point for the LDR group were 0.30 to 0.8 Gy/h, and for the HDR group 1.0 to 3.4 Gy/min. The patients in the HDR group received a total dose of 48-60 Gy (8-10 fractions) during one week. Two fractions were administered per day (at least a 6-h interval). The 3- and 5-year local control rates for patients in the LDR group were 85% and 80%, respectively, and those in the HDR group were both 84%. HDR brachytherapy showed the same lymph-node control rate as did LDR brachytherapy (67% at 5 years). HDR brachytherapy achieved the same locoregional result as did LDR brachytherapy. A converting factor of 0.86 is applicable for HDR in the treatment of early oral tongue cancer. (author)

  4. A study of optimization techniques in HDR brachytherapy for the prostate

    Science.gov (United States)

    Pokharel, Ghana Shyam

    Several studies carried out thus far are in favor of dose escalation to the prostate gland to have better local control of the disease. But optimal way of delivery of higher doses of radiation therapy to the prostate without hurting neighboring critical structures is still debatable. In this study, we proposed that real time high dose rate (HDR) brachytherapy with highly efficient and effective optimization could be an alternative means of precise delivery of such higher doses. This approach of delivery eliminates the critical issues such as treatment setup uncertainties and target localization as in external beam radiation therapy. Likewise, dosimetry in HDR brachytherapy is not influenced by organ edema and potential source migration as in permanent interstitial implants. Moreover, the recent report of radiobiological parameters further strengthen the argument of using hypofractionated HDR brachytherapy for the management of prostate cancer. Firstly, we studied the essential features and requirements of real time HDR brachytherapy treatment planning system. Automating catheter reconstruction with fast editing tools, fast yet accurate dose engine, robust and fast optimization and evaluation engine are some of the essential requirements for such procedures. Moreover, in most of the cases we performed, treatment plan optimization took significant amount of time of overall procedure. So, making treatment plan optimization automatic or semi-automatic with sufficient speed and accuracy was the goal of the remaining part of the project. Secondly, we studied the role of optimization function and constraints in overall quality of optimized plan. We have studied the gradient based deterministic algorithm with dose volume histogram (DVH) and more conventional variance based objective functions for optimization. In this optimization strategy, the relative weight of particular objective in aggregate objective function signifies its importance with respect to other objectives

  5. Intrinsic point-defect balance in self-ion-implanted ZnO.

    Science.gov (United States)

    Neuvonen, Pekka T; Vines, Lasse; Svensson, Bengt G; Kuznetsov, Andrej Yu

    2013-01-04

    The role of excess intrinsic atoms for residual point defect balance has been discriminated by implanting Zn or O ions into Li-containing ZnO and monitoring Li redistribution and electrical resistivity after postimplant anneals. Strongly Li-depleted regions were detected in the Zn-implanted samples at depths beyond the projected range (R(p)) upon annealing ≥ 600 °C, correlating with a resistivity decrease. In contrast, similar anneals of the O-implanted samples resulted in Li accumulation at R(p) and an increased resistivity. Control samples implanted with Ar or Ne ions, yielding similar defect production as for the Zn or O implants but with no surplus of intrinsic atoms, revealed no Li depletion. Thus, the depletion of Li shows evidence of excess Zn interstitials (Zn(I)) being released during annealing of the Zn-implanted samples. These Zn(I)'s convert substitutional Li atoms (Li(Zn)) into highly mobile interstitial ones leading to the strongly Li-depleted regions. In the O-implanted samples, the high resistivity provides evidence of stable O(I)-related acceptors.

  6. Influence of source batch Sk dispersion on dosimetry for prostate cancer treatment with permanent implants

    DEFF Research Database (Denmark)

    Nunez-Cumplido, Eduardo; Perez-Calatayud, J; Casares-Magaz, O

    2015-01-01

    PURPOSE: In clinical practice, specific air kerma strength (SK) value is used in treatment planning system (TPS) permanent brachytherapy implant calculations with (125)I and (103)Pd sources; in fact, commercial TPS provide only one SK input value for all implanted sources and the certified shipment...

  7. Phosphorus-defect interactions during thermal annealing of ion implanted silicon

    Science.gov (United States)

    Keys, Patrick Henry

    Ion implantation of dopant atoms into silicon generates nonequilibrium levels of crystal defects that can lead to the detrimental effects of transient enhanced diffusion (TED), incomplete dopant activation, and p-n junction leakage. In order to control these effects, it is vital to have a clear understanding of dopant-defect interactions and develop models that account for these interactions. This research focuses on experimentally investigating and modeling the clustering of phosphorus dopant atoms with silicon interstitials. Damage recovery of 40keV Si+ implants in phosphorus doped wells is experimentally analyzed. The effects of background phosphorus concentration, self implant dose, and anneal temperature are investigated. Phosphorus concentrations ranging from 2.0 x 1017 to 4.0 x 1019 cm-3 and Si+ doses ranging from 5.0 x 1013 cm-2 to 2.0 x 1014 cm-2 are studied during 650-800°C anneals. A dramatic reduction in the number of interstitials bound in {311} defects with increasing phosphorus background concentration is observed. It is suggested that the reduction of interstitials in {311} defects at high phosphorus concentrations is due to the formation of phosphorus-interstitial clusters (PICs). The critical concentration for clustering (approximately 1.0 x 1019 cm-3 at 750°C) is strongly temperature dependent and in close agreement with the kink concentration of phosphorus diffusion. Information gained from these "well experiments" is applied to the study of direct phosphorus implantation. An experimental study is conducted on 40keV phosphorus implanted to a dose of 1.0 x 1014 cm-2 during 650-800°C anneals. Electrically inactive PICs are shown to form at concentrations below the solid solubility limit due to high interstitial supersaturations. Data useful for developing a model to accurately predict phosphorus diffusion under nonequilibrium conditions are extracted from the experimental results. A cluster-mediated diffusion model is developed using the

  8. Radiation exposure to operating room staff during prostate brachytherapy using iodine-125 seeds; Exposition radiologique de l'equipe operatoire au cours de curietherapies de prostate par implants permanents d'iode-125

    Energy Technology Data Exchange (ETDEWEB)

    Gagna, G.; Amabile, J.C.; Laroche, P. [Service de protection radiologique des armees (SPRA), 1 bis rue du Lieutenant Raoul Batany, 92141 Clamart Cedex (France); Gauron, C. [Institut national de recherche et de securite (INRS), Departement Etudes et Assistance Medicales, 30 rue Olivier Noyer, 75680 Paris Cedex 14 (France)

    2011-04-15

    The French defense radiation protection service (SPRA) and the French national institute for research and safety (INRS) conducted a joint study to assess the radiation exposure to operating room staff during prostate brachytherapy using iodine-125 seeds at the Val-de-Grace military hospital. The purpose of the study was the assessment of the effective doses, the equivalent doses to the extremities and lens received by a novice team, the different ambient dose equivalent rates measurements and the delineation of areas. After six brachy-therapies, all the recorded doses with whole-body InLight{sup R} OSL and nanoDot{sup R} dosimeters remained below the detection limit for the whole staff. The dose rate measured at the end of implantation by an AT1123{sup R} survey meter is about 170 {mu}Sv/h at the perineum of the patient. The controlled area limit is estimated to be about 20 cm from the patient perineum. From these results, the authors propose recommendations for the categorization of workers, the delineation of areas and the dose monitoring procedures. This study demonstrates that real-time ultrasound-guided trans-perineal prostate brachytherapy delivers low dose to the operators because of the radioactive source characteristics and the instrumentation providing an effective radiation protection for the surgical team. (authors)

  9. 1251 seed calibration using afterloading equipment SeedSelectron. Practical solution to meet the recommendations of the AAPM

    International Nuclear Information System (INIS)

    Perez-Calatayud, J.; Richart, J.; Perez-Garcia, J.; Guirado, D.; Ballester, F.; Rodriguez, S.; Santos, M.; Depiaggio, M.; Carmona, V.; Lliso, F.; Camacho, C.; Pujades, M. C.

    2011-01-01

    SeedSelectron is a system used in the afterloader permanent implant brachytherapy seeds 1-125 interstitial prostate. Two aspects are critical when you can meet the recommendations of the AAPM: a practical difficulty to check the quantity of seed required, and the great uncertainty of all measured diodes. The purpose of this paper is to present a practical solution that has been adopted to implement the recommendations of the AAPM

  10. 1251 seed calibration using afterloading equipment Seed Selectron. Practical solution to meet the recommendations of the AAPM; Calibracion de semillas de {sup 1}25I usando el equipo de carga difereida SeedSelectron. Solucion practica para cumplir las recomendaciones de la AAPM

    Energy Technology Data Exchange (ETDEWEB)

    Perez-Calatayud, J.; Richart, J.; Perez-Garcia, J.; Guirado, D.; Ballester, F.; Rodriguez, S.; Santos, M.; Depiaggio, M.; Carmona, V.; Lliso, F.; Camacho, C.; Pujades, M. C.

    2011-07-01

    Seed Selectron is a system used in the after loader permanent implant brachytherapy seeds 1-125 interstitial prostate. Two aspects are critical when you can meet the recommendations of the AAPM: a practical difficulty to check the quantity of seed required, and the great uncertainty of all measured diodes. The purpose of this paper is to present a practical solution that has been adopted to implement the recommendations of the AAPM.

  11. Brachytherapy Application With In Situ Dose Painting Administered by Gold Nanoparticle Eluters

    International Nuclear Information System (INIS)

    Sinha, Neeharika; Cifter, Gizem; Sajo, Erno; Kumar, Rajiv; Sridhar, Srinivas; Nguyen, Paul L.; Cormack, Robert A.; Makrigiorgos, G. Mike; Ngwa, Wilfred

    2015-01-01

    Purpose: Recent studies show promise that administering gold nanoparticles (GNP) to tumor cells during brachytherapy could significantly enhance radiation damage to the tumor. A new strategy proposed for sustained administration of the GNP in prostate tumors is to load them into routinely used brachytherapy spacers for customizable in situ release after implantation. This in silico study investigated the intratumor biodistribution and corresponding dose enhancement over time due to GNP released from such GNP-loaded brachytherapy spacers (GBS). Method and Materials: An experimentally determined intratumoral diffusion coefficient (D) for 10-nm nanoparticles was used to estimate D for other sizes by using the Stokes-Einstein equation. GNP concentration profiles, obtained using D, were then used to calculate the corresponding dose enhancement factor (DEF) for each tumor voxel, using dose painting-by-numbers approach, for times relevant to the considered brachytherapy sources' lifetimes. The investigation was carried out as a function of GNP size for the clinically applicable low-dose-rate brachytherapy sources iodine-125 (I-125), palladium-103 (Pd-103), and cesium-131 (Cs-131). Results: Results showed that dose enhancement to tumor voxels and subvolumes during brachytherapy can be customized by varying the size of GNP released or eluted from the GBS. For example, using a concentration of 7 mg/g GNP, significant DEF (>20%) could be achieved 5 mm from a GBS after 5, 12, 25, 46, 72, 120, and 195 days, respectively, for GNP sizes of 2, 5, 10, 20, 30, and 50 nm and for 80 nm when treating with I-125. Conclusions: Analyses showed that using Cs-131 provides the highest dose enhancement to tumor voxels. However, given its relatively longer half-life, I-125 presents the most flexibility for customizing the dose enhancement as a function of GNP size. These findings provide a useful reference for further work toward development of potential new brachytherapy application

  12. Brachytherapy Application With In Situ Dose Painting Administered by Gold Nanoparticle Eluters

    Energy Technology Data Exchange (ETDEWEB)

    Sinha, Neeharika [Department of Sciences, Wentworth Institute of Technology, Boston, Massachusetts (United States); Cifter, Gizem [Department of Physics and Applied Physics, University of Massachusetts, Lowell, Massachusetts (United States); Department of Radiation Oncology, Dana-Farber Cancer Institute, Brigham and Women' s Hospital and Harvard Medical School, Boston, Massachusetts (United States); Sajo, Erno [Department of Physics and Applied Physics, University of Massachusetts, Lowell, Massachusetts (United States); Kumar, Rajiv; Sridhar, Srinivas [Department of Radiation Oncology, Dana-Farber Cancer Institute, Brigham and Women' s Hospital and Harvard Medical School, Boston, Massachusetts (United States); Electronic Materials Research Institute and Department of Physics, Northeastern University, Boston, Massachusetts (United States); Nguyen, Paul L.; Cormack, Robert A.; Makrigiorgos, G. Mike [Department of Radiation Oncology, Dana-Farber Cancer Institute, Brigham and Women' s Hospital and Harvard Medical School, Boston, Massachusetts (United States); Ngwa, Wilfred, E-mail: wngwa@lroc.harvard.edu [Department of Physics and Applied Physics, University of Massachusetts, Lowell, Massachusetts (United States); Department of Radiation Oncology, Dana-Farber Cancer Institute, Brigham and Women' s Hospital and Harvard Medical School, Boston, Massachusetts (United States)

    2015-02-01

    Purpose: Recent studies show promise that administering gold nanoparticles (GNP) to tumor cells during brachytherapy could significantly enhance radiation damage to the tumor. A new strategy proposed for sustained administration of the GNP in prostate tumors is to load them into routinely used brachytherapy spacers for customizable in situ release after implantation. This in silico study investigated the intratumor biodistribution and corresponding dose enhancement over time due to GNP released from such GNP-loaded brachytherapy spacers (GBS). Method and Materials: An experimentally determined intratumoral diffusion coefficient (D) for 10-nm nanoparticles was used to estimate D for other sizes by using the Stokes-Einstein equation. GNP concentration profiles, obtained using D, were then used to calculate the corresponding dose enhancement factor (DEF) for each tumor voxel, using dose painting-by-numbers approach, for times relevant to the considered brachytherapy sources' lifetimes. The investigation was carried out as a function of GNP size for the clinically applicable low-dose-rate brachytherapy sources iodine-125 (I-125), palladium-103 (Pd-103), and cesium-131 (Cs-131). Results: Results showed that dose enhancement to tumor voxels and subvolumes during brachytherapy can be customized by varying the size of GNP released or eluted from the GBS. For example, using a concentration of 7 mg/g GNP, significant DEF (>20%) could be achieved 5 mm from a GBS after 5, 12, 25, 46, 72, 120, and 195 days, respectively, for GNP sizes of 2, 5, 10, 20, 30, and 50 nm and for 80 nm when treating with I-125. Conclusions: Analyses showed that using Cs-131 provides the highest dose enhancement to tumor voxels. However, given its relatively longer half-life, I-125 presents the most flexibility for customizing the dose enhancement as a function of GNP size. These findings provide a useful reference for further work toward development of potential new brachytherapy application

  13. Dosimetric properties of a novel brachytherapy balloon applicator for the treatment of malignant brain-tumor resection-cavity margins

    International Nuclear Information System (INIS)

    Dempsey, James F.; Williams, Jeffery A.; Stubbs, James B.; Patrick, Timothy J.; Williamson, Jeffrey F.

    1998-01-01

    Purpose: This paper characterizes the dosimetric properties of a novel balloon brachytherapy applicator for the treatment of the tissue surrounding the resection cavity of a malignant brain tumor. Methods and Materials: The applicator consists of an inflatable silicone balloon reservoir attached to a positionable catheter that is intraoperatively implanted into the resection cavity and postoperatively filled with a liquid radionuclide solution. A simple dosimetric model, valid in homogeneous media and based on results from Monte Carlo photon-transport simulations, was used to determine the dosimetric characteristics of spherical geometry balloons filled with photon-emitting radionuclide solutions. Fractional depth-dose (FDD) profiles, along with activity densities, and total activities needed to achieve specified dose rates were studied as a function of photon energy and source-containment geometry. Dose-volume histograms (DVHs) were calculated to compare idealized balloon-applicator treatments to conventional 125 I seed volume implants. Results: For achievable activity densities and total activities, classical low dose rate (LDR) treatments of residual disease at distances of up to 1 cm from the resection cavity wall are possible with balloon applicators having radii between 0.5 cm and 2.5 cm. The dose penetration of these applicators increases approximately linearly with balloon radius. The FDD profile can be made significantly more or less penetrating by combining selection of radionuclide with source-geometry manipulation. Comparisons with 125 I seed-implant DVHs show that the applicator can provide a more conformal therapy with no target tissue underdosing, less target tissue overdosing, and no healthy tissue ''hot spots;'' however, more healthy tissue volume receives a dose of the prescribed dosage or less. Conclusions: This device, when filled with 125 I solution, is suitable for classical LDR treatments and may be preferable to 125 I interstitial

  14. Optical fibre luminescence sensor for real-time LDR brachytherapy dosimetry

    Science.gov (United States)

    Woulfe, P.; O'Keeffe, S.; Sullivan, F. J.

    2018-02-01

    An optical fibre sensor for monitoring low dose radiation is presented. The sensor is based on a scintillation material embedded within the optical fibre core, which emits visible light when exposed to low level ionising radiation. The incident level of ionising radiation can be determined by analysing the optical emission. An optical fibre sensor is developed, based on radioluminescence whereby radiation sensitive scintillation material, terbium doped gadolinium oxysulphide (Gd2O2S:Tb), is embedded in a cavity of 700μm of a 1mm plastic optical fibre. The sensor is designed for in-vivo monitoring of the radiation dose during radio-active seed implantation for low dose rate (LDR) brachytherapy, in prostate cancer treatment, providing radiation oncologists with real-time information of the radiation dose to the target area and/or nearby organs at risk (OARs). The radiation from the brachytherapy seeds causes emission of visible light from the scintillation material through the process of radioluminescence, which penetrates the fibre, propagating along the optical fibre for remote detection using a multi-pixel photon counter. The sensor demonstrates a high sensitivity to 0.397mCi of Iodine125, the radioactive source most commonly used in brachytherapy for treating prostate cancer.

  15. Novel use of the Contura for high dose rate cranial brachytherapy.

    Science.gov (United States)

    Scanderbeg, Daniel J; Alksne, John F; Lawson, Joshua D; Murphy, Kevin T

    2011-01-01

    A popular choice for treatment of recurrent gliomas was cranial brachytherapy using the GliaSite Radiation Therapy System. However, this device was taken off the market in late 2008, thus leaving a treatment void. This case study presents our experience treating a cranial lesion for the first time using a Contura multilumen, high-dose-rate (HDR) brachytherapy balloon applicator. The patient was a 47-year-old male who was diagnosed with a recurrent right frontal anaplastic oligodendroglioma. Previous radiosurgery made him a good candidate for brachytherapy. An intracavitary HDR balloon brachytherapy device (Contura) was placed in the resection cavity and treated with a single fraction of 20 Gy. The implant, treatment, and removal of the device were all completed without incident. Dosimetry of the device was excellent because the dose conformed very well to the target. V90, V100, V150, and V200 were 98.9%, 95.7%, 27.2, and 8.8 cc, respectively. This patient was treated successfully using the Contura multilumen balloon. Contura was originally designed for deployment in a postlumpectomy breast for treatment by accelerated partial breast irradiation. Being an intracavitary balloon device, its similarity to the GliaSite system makes it a viable replacement candidate. Multiple lumens in the device also make it possible to shape the dose delivered to the target, something not possible before with the GliaSite applicator. Copyright © 2011 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

  16. A Phase III Randomized Trial of the Timing of Meloxicam With Iodine-125 Prostate Brachytherapy

    International Nuclear Information System (INIS)

    Crook, Juanita; Patil, Nikhilesh; Wallace, Kris; Borg, Jette; Zhou, David; Ma, Clement; Pond, Greg

    2010-01-01

    Purpose: Nonsteroidal anti-inflammatory medication is used to reduce prostate edema and urinary symptoms following prostate brachytherapy. We hypothesized that a cyclooxygenase-2 (COX-2) inhibitor regimen started 1 week prior to seed implant might diminish the inflammatory response, thus reducing edema, retention rates, and symptom severity. Methods and Materials: From March 2004 to February 2008, 316 men consented to an institutional review board-approved randomized study of a 4-week course of meloxicam, 7.5 mg orally twice per day, starting either on the day of implant or 1 week prior to implant. Brachytherapy was performed using iodine-125 seeds and was preplanned and performed under transrectal ultrasound (TRUS) and fluoroscopic guidance. Prostate volume obtained by MR imaging at 1 month was compared to baseline prostate volume obtained by TRUS planimetry and expressed as an edema factor. The trial endpoints were prostate edema at 1 month, International Prostate Symptom Score (IPSS) questionnaire results at 1 and 3 months, and any need for catheterization. Results: Results for 300 men were analyzed. Median age was 61 (range, 45-79 years), and median TRUS prostate volume was 35.7 cc (range, 18.1-69.5 cc). Median IPSS at baseline was 5 (range, 0-24) and was 15 at 1 month, 16 at 3 months, and 10 at 6 months. Catheterization was required for 7% of patients (6.2% day 0 arm vs. 7.9% day -7 arm; p = 0.65). The median edema factor at 1 month was 1.02 (range, 0.73-1.7). 1.01 day 0 arm vs. 1.05 day -7 arm. Baseline prostate volume remained the primary predictor of postimplant urinary retention. Conclusions: Starting meloxicam 1 week prior to brachytherapy compared to starting immediately after the procedure did not reduce 1-month edema, improve IPSSs at 1 or 3 months, or reduce the need for catheterization.

  17. High dose rate versus low dose rate brachytherapy in the treatment of stage IIIB cervical cancer, and the importance of brachytherapy timing

    International Nuclear Information System (INIS)

    Petereit, Daniel G.; Sarkaria, Jann N.; Czyzewski, Ann; Buchler, Dolores A.

    1996-01-01

    Purpose/Objective: To determine the efficacy of HDR versus LDR brachytherapy for Stage IIIB cervical cancer patients. Material and Methods: Forty-four HDR patients were retrospectively compared to 51 LDR patients treated at the same institution from 1977 to 1988 (LDR) and from 1989 to 1995 (HDR). A tumor burden score (TBS) of 2-9 was calculated for both groups of patients to assess volume of disease (2-4 low tumor burden, 5-9 high tumor burden). LDR and HDR patients received 60 Gy to the whole pelvis at 1.7 Gy/Fx. The majority of LDR patients were treated after completion of external beam radiation (EBR) with one 25 Gy implant to Point A (55 cGy/h). The HDR patients were treated with 4-5 HDR fractions of 3.7 Gy to 5.8 Gy/Fx for an LDR equivalence of 20-25 Gy (median dose/Fx 4.3 Gy, median insertion number 5). Clinical endpoints were calculated actuarially with significance determined by the log rank test and the relative risk (RR). Results: The median follow-up for the HDR and LDR groups was 1.8 and 5 years, respectively. Pelvic control and survival was better in the LDR group than the HDR group, 51%, 73%, 32%, 44% (p = 0.004, RR = 0.4), with grade III and above RTOG complications of 19% and 15%, respectively. The median age and performance status were similar between the two groups; however, a TBS score ≥7 was present in 23% of the HDR patients and in 9% of the LDR patients. Pelvic control in the HDR group was 58% with a TBS ≤4, and 17% with a TBS >4 (p = 0.01, RR = 0.4). The median EBR dose at the first HDR insertion was 31 Gy while all the LDR patients received 60 Gy before the insertion. Pelvic control rates in Table 1 indicate a trend towards improved outcome within the HDR group and same TBS if more external beam radiation was given prior to the first HDR insertion. Pelvic control was also higher within the HDR group when Point A received a BED Gy 10≥100 versus <100: 62%, 40%, respectively (RR 0.6). Conclusion: These retrospective results of HDR versus

  18. Comparative studies on permanent prostate brachytherapy: pre-plan and real-time transrectal ultrasound guided iodine-125 seed implants at Korle-Bu Teaching Hospital, Ghana

    International Nuclear Information System (INIS)

    Kalolo, L.T.

    2013-06-01

    This research was carried out to investigate and compare the real-time and pre-plan implant at the Radiotherapy Department of the Korle Bu Teaching Hospital, Ghana. Prowess Panther 4.5 treatment planning system and variseed 7.2 software were used for pre-plan and real-time implant respectively. The study was conducted for eighty three (83) patients treated for prostate cancer through real-time implant brachytherapy between september, 2008 to April, 2013. Thirty one patients (31) patients whose ultrasound images were available were selected for the pre-plan study. The slices of ultrasound images were re-drawn on transparent A-4 sheets and later on scanned, contoured and registered in the treatment planning system (prowess 4.5). After planning, the volume to be implanted, total number of needles, seeds and the total activity of the source were displayed. Comparison was done withe the pre-plan and real-time implant. In both cases the variation was below 5% as recommended in dosimetry. About 30% - 40% of the imported seeds were left un-used due to over-estimation of seeds ordered from the manufacturer (BARD Company-USA). Hence this work (pre-plan) aims to solve this problem. The comparison for dosimetric parameters was assessed for prostate, urethra and rectum as (V 95%, V 100%, V 150%, D90Gy, D90%), (D90Gy, D90%, D30Gy, D30% ) and (V 100%, D30Gy and D30%) respectively and the variation were within the limit of ± 5%. Comparison of dosimetric values for this work were done with other institutions, like Karolinska university hospital, Sweden, The institute of Curie/ hospital Cochin Group Paris-France and European recommendations. The values reported at Korle - Bu teaching hospital (this work) were in good agreement with the international guidelines. (au)

  19. SU-G-201-15: Nomogram as an Efficient Dosimetric Verification Tool in HDR Prostate Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Liang, J; Todor, D [Virginia Commonwealth University, Richmond, VA (United States)

    2016-06-15

    Purpose: Nomogram as a simple QA tool for HDR prostate brachytherapy treatment planning has been developed and validated clinically. Reproducibility including patient-to-patient and physician-to-physician variability was assessed. Methods: The study was performed on HDR prostate implants from physician A (n=34) and B (n=15) using different implant techniques and planning methodologies. A nomogram was implemented as an independent QA of computer-based treatment planning before plan execution. Normalized implant strength (total air kerma strength Sk*t in cGy cm{sup 2} divided by prescribed dose in cGy) was plotted as a function of PTV volume and total V100. A quadratic equation was used to fit the data with R{sup 2} denoting the model predictive power. Results: All plans showed good target coverage while OARs met the dose constraint guidelines. Vastly different implant and planning styles were reflected on conformity index (entire dose matrix V100/PTV volume, physician A implants: 1.27±0.14, physician B: 1.47±0.17) and PTV V150/PTV volume ratio (physician A: 0.34±0.09, physician B: 0.24±0.07). The quadratic model provided a better fit for the curved relationship between normalized implant strength and total V100 (or PTV volume) than a simple linear function. Unlike the normalized implant strength versus PTV volume nomogram which differed between physicians, a unique quadratic model based nomogram (Sk*t)/D=−0.0008V2+0.0542V+1.1185 (R{sup 2}=0.9977) described the dependence of normalized implant strength on total V100 over all the patients from both physicians despite two different implant and planning philosophies. Normalized implant strength - total V100 model also generated less deviant points distorting the smoothed ones with a significantly higher correlation. Conclusion: A simple and universal, excel-based nomogram was created as an independent calculation tool for HDR prostate brachytherapy. Unlike similar attempts, our nomogram is insensitive to implant

  20. WE-FG-BRA-02: Docetaxel Eluting Brachytherapy Spacers for Local Chemo-Radiation Therapy in Prostate Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Belz, J [Northeastern University, Boston, MA (United States); Kumar, R; Sridhar, S [Northeastern University & Dana Farber Cancer Institute, Boston, MA (United States); Makrigiorgos, G; Nguyen, P [Dana Farber Cancer Institute, Boston, MA (United States); D’Amico, A [Brigham & Women’s Hospital, Boston, MA (United States); Cormack, R [Harvard Medical School, Boston, MA (United States)

    2016-06-15

    Purpose: We propose an innovative combinatorial treatment strategy of Local ChemoRadiation Therapy (LCRT) using a sustained drug delivery platform in the form of a spacer to locally radio-sensitize the prostate with Docetaxel (DTX) enabling a synergistic cure with the use of lower radiation doses. These biodegradable spacers are physically similar to the inert spacers routinely used in prostate brachytherapy but are now loaded with formulations of DTX. Methods: Spacers were loaded with ∼500µg Docetaxel (DTX) for prostate cancer studies. The implants were characterized in vitro using SEM and HPLC. The release kinetic studies were carried out in buffer (pH 6.0) at 37°C. Subcutaneous PC3 tumors were xenografted in nude mice. Prostate cancer studies were done with and without radiation using SARRP at 5Gy, 10Gy, and 15Gy. Drug-loaded implants were injected once intratumorally using an 18G brachytherapy needle. Results: The release study in vitro showed a highly sustained release for multiple weeks at therapeutically relevant doses. The monotherapy with local DTX spacer showed sustained tumor inhibition compared to empty implants and an equivalent DTX dose given systemically. At 40 days, 89% survival was observed for mice treated with DTX implants compared with 0% in all other treatment groups. The combined treatment with local DTX spacer and radiation (10Gy) showed the highest degree of tumor suppression (significant tumor growth inhibition by day 90). The control mice showed continuous tumor growth and were scarified by day 56. Groups of mice treated with DTX-spacer or radiation alone showed initial tumor suppression but growth continued after day 60. A larger experiment is ongoing. Conclusion: This approach provides localized delivery of the chemotherapeutic sensitizer directly to the tumor and avoids the toxicities associated with both brachytherapy and current systemic delivery of docetaxel. Sustained release of DTX is an effective chemotherapy option alone or