WorldWideScience

Sample records for epid cine images

  1. Detection and correction for EPID and gantry sag during arc delivery using cine EPID imaging.

    Science.gov (United States)

    Rowshanfarzad, Pejman; Sabet, Mahsheed; O'Connor, Daryl J; McCowan, Peter M; McCurdy, Boyd M C; Greer, Peter B

    2012-02-01

    Electronic portal imaging devices (EPIDs) have been studied and used for pretreatment and in-vivo dosimetry applications for many years. The application of EPIDs for dosimetry in arc treatments requires accurate characterization of the mechanical sag of the EPID and gantry during rotation. Several studies have investigated the effects of gravity on the sag of these systems but each have limitations. In this study, an easy experiment setup and accurate algorithm have been introduced to characterize and correct for the effect of EPID and gantry sag during arc delivery. Three metallic ball bearings were used as markers in the beam: two of them fixed to the gantry head and the third positioned at the isocenter. EPID images were acquired during a 360° gantry rotation in cine imaging mode. The markers were tracked in EPID images and a robust in-house developed MATLAB code was used to analyse the images and find the EPID sag in three directions as well as the EPID + gantry sag by comparison to the reference gantry zero image. The algorithm results were then tested against independent methods. The method was applied to compare the effect in clockwise and counter clockwise gantry rotations and different source-to-detector distances (SDDs). The results were monitored for one linear accelerator over a course of 15 months and six other linear-accelerators from two treatment centers were also investigated using this method. The generalized shift patterns were derived from the data and used in an image registration algorithm to correct for the effect of the mechanical sag in the system. The Gamma evaluation (3%, 3 mm) technique was used to investigate the improvement in alignment of cine EPID images of a fixed field, by comparing both individual images and the sum of images in a series with the reference gantry zero image. The mechanical sag during gantry rotation was dependent on the gantry angle and was larger in the in-plane direction, although the patterns were not

  2. Verification of the linac isocenter for stereotactic radiosurgery using cine-EPID imaging and arc delivery

    International Nuclear Information System (INIS)

    Rowshanfarzad, Pejman; Sabet, Mahsheed; O' Connor, Daryl J.; Greer, Peter B.

    2011-01-01

    Purpose:Verification of the mechanical isocenter position is required as part of comprehensive quality assurance programs for stereotactic radiosurgery/radiotherapy (SRS/SRT) treatments. Several techniques have been proposed for this purpose but each of them has certain drawbacks. In this paper, a new efficient and more comprehensive method using cine-EPID images has been introduced for automatic verification of the isocenter with sufficient accuracy for stereotactic applications. Methods: Using a circular collimator fixed to the gantry head to define the field, EPID images of a Winston-Lutz phantom were acquired in cine-imaging mode during 360 deg. gantry rotations. A robust matlab code was developed to analyze the data by finding the center of the field and the center of the ball bearing shadow in each image with sub-pixel accuracy. The distance between these two centers was determined for every image. The method was evaluated by comparison to results of a mechanical pointer and also by detection of a manual shift applied to the phantom position. The repeatability and reproducibility of the method were tested and it was also applied to detect couch and collimator wobble during rotation. Results:The accuracy of the algorithm was 0.03 ± 0.02 mm. The repeatability was less than 3 μm and the reproducibility was less than 86 μm. The time elapsed for the analysis of more than 100 cine images of Varian aS1000 and aS500 EPIDs were ∼65 and 20 s, respectively. Processing of images taken in integrated mode took 0.1 s. The output of the analysis software is printable and shows the isocenter shifts as a function of angle in both in-plane and cross-plane directions. It gives warning messages where the shifts exceed the criteria for SRS/SRT and provides useful data for the necessary adjustments in the system including bearing system and/or room lasers. Conclusions: The comprehensive method introduced in this study uses cine-images, is highly accurate, fast, and independent

  3. Verification of the linac isocenter for stereotactic radiosurgery using cine-EPID imaging and arc delivery

    Energy Technology Data Exchange (ETDEWEB)

    Rowshanfarzad, Pejman; Sabet, Mahsheed; O' Connor, Daryl J.; Greer, Peter B. [School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, New South Wales 2308 (Australia); Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, New South Wales 2310, Australia and School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, New South Wales 2308 (Australia)

    2011-07-15

    Purpose:Verification of the mechanical isocenter position is required as part of comprehensive quality assurance programs for stereotactic radiosurgery/radiotherapy (SRS/SRT) treatments. Several techniques have been proposed for this purpose but each of them has certain drawbacks. In this paper, a new efficient and more comprehensive method using cine-EPID images has been introduced for automatic verification of the isocenter with sufficient accuracy for stereotactic applications. Methods: Using a circular collimator fixed to the gantry head to define the field, EPID images of a Winston-Lutz phantom were acquired in cine-imaging mode during 360 deg. gantry rotations. A robust matlab code was developed to analyze the data by finding the center of the field and the center of the ball bearing shadow in each image with sub-pixel accuracy. The distance between these two centers was determined for every image. The method was evaluated by comparison to results of a mechanical pointer and also by detection of a manual shift applied to the phantom position. The repeatability and reproducibility of the method were tested and it was also applied to detect couch and collimator wobble during rotation. Results:The accuracy of the algorithm was 0.03 {+-} 0.02 mm. The repeatability was less than 3 {mu}m and the reproducibility was less than 86 {mu}m. The time elapsed for the analysis of more than 100 cine images of Varian aS1000 and aS500 EPIDs were {approx}65 and 20 s, respectively. Processing of images taken in integrated mode took 0.1 s. The output of the analysis software is printable and shows the isocenter shifts as a function of angle in both in-plane and cross-plane directions. It gives warning messages where the shifts exceed the criteria for SRS/SRT and provides useful data for the necessary adjustments in the system including bearing system and/or room lasers. Conclusions: The comprehensive method introduced in this study uses cine-images, is highly accurate, fast, and

  4. The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking

    Energy Technology Data Exchange (ETDEWEB)

    Yip, Stephen, E-mail: syip@lroc.harvard.edu; Rottmann, Joerg; Berbeco, Ross [Department of Radiation Oncology, Brigham and Women' s Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts 02115 (United States)

    2014-06-15

    Purpose: Although reduction of the cine electronic portal imaging device (EPID) acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft-tissue motion leading to poor autotracking results. The impact of motion blurring and image noise on the tracking performance was investigated. Methods: Phantom and patient images were acquired at a frame rate of 12.87 Hz with an amorphous silicon portal imager (AS1000, Varian Medical Systems, Palo Alto, CA). The maximum frame rate of 12.87 Hz is imposed by the EPID. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for autotracking. The difference between the programmed and autotracked positions of a Las Vegas phantom moving in the superior-inferior direction defined the tracking error (δ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at 11 field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise are correlated with δ using Pearson correlation coefficient (R). Results: For both phantom and patient studies, the autotracking errors increased at frame rates lower than 4.29 Hz. Above 4.29 Hz, changes in errors were negligible withδ < 1.60 mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R = 0.94) and patient studies (R = 0.72). Moderate to poor correlation was found between image noise and tracking error with R −0.58 and −0.19 for both studies, respectively. Conclusions: Cine EPID

  5. The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking

    International Nuclear Information System (INIS)

    Yip, Stephen; Rottmann, Joerg; Berbeco, Ross

    2014-01-01

    Purpose: Although reduction of the cine electronic portal imaging device (EPID) acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft-tissue motion leading to poor autotracking results. The impact of motion blurring and image noise on the tracking performance was investigated. Methods: Phantom and patient images were acquired at a frame rate of 12.87 Hz with an amorphous silicon portal imager (AS1000, Varian Medical Systems, Palo Alto, CA). The maximum frame rate of 12.87 Hz is imposed by the EPID. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for autotracking. The difference between the programmed and autotracked positions of a Las Vegas phantom moving in the superior-inferior direction defined the tracking error (δ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at 11 field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise are correlated with δ using Pearson correlation coefficient (R). Results: For both phantom and patient studies, the autotracking errors increased at frame rates lower than 4.29 Hz. Above 4.29 Hz, changes in errors were negligible withδ < 1.60 mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R = 0.94) and patient studies (R = 0.72). Moderate to poor correlation was found between image noise and tracking error with R −0.58 and −0.19 for both studies, respectively. Conclusions: Cine EPID

  6. Computerized method for estimation of the location of a lung tumor on EPID cine images without implanted markers in stereotactic body radiotherapy

    International Nuclear Information System (INIS)

    Arimura, H; Toyofuku, F; Higashida, Y; Onizuka, Y; Terashima, H; Egashira, Y; Shioyama, Y; Nomoto, S; Honda, H; Nakamura, K; Yoshidome, S; Anai, S

    2009-01-01

    The purpose of this study was to develop a computerized method for estimation of the location of a lung tumor in cine images on an electronic portal imaging device (EPID) without implanted markers during stereotactic body radiotherapy (SBRT). Each tumor region was segmented in the first EPID cine image, i.e., reference portal image, based on a multiple-gray level thresholding technique and a region growing technique, and then the image including the tumor region was cropped as a 'tumor template' image. The tumor location was determined as the position in which the tumor template image took the maximum cross-correlation value within each consecutive portal image, which was acquired in cine mode on the EPID in treatment. EPID images with 512 x 384 pixels (pixel size: 0.56 mm) were acquired at a sampling rate of 0.5 frame s -1 by using energies of 4, 6 or 10 MV on linear accelerators. We applied our proposed method to EPID cine images (226 frames) of 12 clinical cases (ages: 51-83, mean: 72) with a non-small cell lung cancer. As a result, the average location error between tumor points obtained by our method and the manual method was 1.47 ± 0.60 mm. This preliminary study suggests that our method based on the tumor template matching technique might be feasible for tracking the location of a lung tumor without implanted markers in SBRT.

  7. An initial study on the estimation of time-varying volumetric treatment images and 3D tumor localization from single MV cine EPID images

    Energy Technology Data Exchange (ETDEWEB)

    Mishra, Pankaj, E-mail: pankaj.mishra@varian.com; Mak, Raymond H.; Rottmann, Joerg; Bryant, Jonathan H.; Williams, Christopher L.; Berbeco, Ross I.; Lewis, John H. [Brigham and Women' s Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts 02115 (United States); Li, Ruijiang [Department of Radiation Oncology, Stanford University School of Medicine, Stanford, California 94305 (United States)

    2014-08-15

    Purpose: In this work the authors develop and investigate the feasibility of a method to estimate time-varying volumetric images from individual MV cine electronic portal image device (EPID) images. Methods: The authors adopt a two-step approach to time-varying volumetric image estimation from a single cine EPID image. In the first step, a patient-specific motion model is constructed from 4DCT. In the second step, parameters in the motion model are tuned according to the information in the EPID image. The patient-specific motion model is based on a compact representation of lung motion represented in displacement vector fields (DVFs). DVFs are calculated through deformable image registration (DIR) of a reference 4DCT phase image (typically peak-exhale) to a set of 4DCT images corresponding to different phases of a breathing cycle. The salient characteristics in the DVFs are captured in a compact representation through principal component analysis (PCA). PCA decouples the spatial and temporal components of the DVFs. Spatial information is represented in eigenvectors and the temporal information is represented by eigen-coefficients. To generate a new volumetric image, the eigen-coefficients are updated via cost function optimization based on digitally reconstructed radiographs and projection images. The updated eigen-coefficients are then multiplied with the eigenvectors to obtain updated DVFs that, in turn, give the volumetric image corresponding to the cine EPID image. Results: The algorithm was tested on (1) Eight digital eXtended CArdiac-Torso phantom datasets based on different irregular patient breathing patterns and (2) patient cine EPID images acquired during SBRT treatments. The root-mean-squared tumor localization error is (0.73 ± 0.63 mm) for the XCAT data and (0.90 ± 0.65 mm) for the patient data. Conclusions: The authors introduced a novel method of estimating volumetric time-varying images from single cine EPID images and a PCA-based lung motion model

  8. SU-E-J-112: The Impact of Cine EPID Image Acquisition Frame Rate On Markerless Soft-Tissue Tracking

    Energy Technology Data Exchange (ETDEWEB)

    Yip, S; Rottmann, J; Berbeco, R [Brigham and Women' s Hospital, Boston, MA (United States)

    2014-06-01

    Purpose: Although reduction of the cine EPID acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft-tissue motion leading to poor auto-tracking results. The impact of motion blurring and image noise on the tracking performance was investigated. Methods: Phantom and patient images were acquired at a frame rate of 12.87Hz on an AS1000 portal imager. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for auto-tracking. The difference between the programmed and auto-tracked positions of a Las Vegas phantom moving in the superior-inferior direction defined the tracking error (δ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at eleven field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise were correlated with δ using Pearson correlation coefficient (R). Results: For both phantom and patient studies, the auto-tracking errors increased at frame rates lower than 4.29Hz. Above 4.29Hz, changes in errors were negligible with δ<1.60mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R=0.94) and patient studies (R=0.72). Moderate to poor correlation was found between image noise and tracking error with R -0.58 and -0.19 for both studies, respectively. Conclusion: An image acquisition frame rate of at least 4.29Hz is recommended for cine EPID tracking. Motion blurring in images with frame rates below 4.39Hz can substantially reduce the

  9. SU-E-J-112: The Impact of Cine EPID Image Acquisition Frame Rate On Markerless Soft-Tissue Tracking

    International Nuclear Information System (INIS)

    Yip, S; Rottmann, J; Berbeco, R

    2014-01-01

    Purpose: Although reduction of the cine EPID acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft-tissue motion leading to poor auto-tracking results. The impact of motion blurring and image noise on the tracking performance was investigated. Methods: Phantom and patient images were acquired at a frame rate of 12.87Hz on an AS1000 portal imager. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for auto-tracking. The difference between the programmed and auto-tracked positions of a Las Vegas phantom moving in the superior-inferior direction defined the tracking error (δ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at eleven field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise were correlated with δ using Pearson correlation coefficient (R). Results: For both phantom and patient studies, the auto-tracking errors increased at frame rates lower than 4.29Hz. Above 4.29Hz, changes in errors were negligible with δ<1.60mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R=0.94) and patient studies (R=0.72). Moderate to poor correlation was found between image noise and tracking error with R -0.58 and -0.19 for both studies, respectively. Conclusion: An image acquisition frame rate of at least 4.29Hz is recommended for cine EPID tracking. Motion blurring in images with frame rates below 4.39Hz can substantially reduce the

  10. Verification of respiratory-gated radiotherapy with new real-time tumour-tracking radiotherapy system using cine EPID images and a log file

    Science.gov (United States)

    Shiinoki, Takehiro; Hanazawa, Hideki; Yuasa, Yuki; Fujimoto, Koya; Uehara, Takuya; Shibuya, Keiko

    2017-02-01

    A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were  ⩽3 mm and  ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases. This work was partly presented at the 58th Annual meeting of American Association of Physicists in Medicine.

  11. Verification of respiratory-gated radiotherapy with new real-time tumour-tracking radiotherapy system using cine EPID images and a log file.

    Science.gov (United States)

    Shiinoki, Takehiro; Hanazawa, Hideki; Yuasa, Yuki; Fujimoto, Koya; Uehara, Takuya; Shibuya, Keiko

    2017-02-21

    A combined system comprising the TrueBeam linear accelerator and a new real-time tumour-tracking radiotherapy system, SyncTraX, was installed at our institution. The objectives of this study are to develop a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine electronic portal image device (EPID) images and a log file and to verify this treatment in clinical cases. Respiratory-gated radiotherapy was performed using TrueBeam and the SyncTraX system. Cine EPID images and a log file were acquired for a phantom and three patients during the course of the treatment. Digitally reconstructed radiographs (DRRs) were created for each treatment beam using a planning CT set. The cine EPID images, log file, and DRRs were analysed using a developed software. For the phantom case, the accuracy of the proposed method was evaluated to verify the respiratory-gated radiotherapy. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker used as an internal surrogate were calculated to evaluate the gating accuracy and set-up uncertainty in the superior-inferior (SI), anterior-posterior (AP), and left-right (LR) directions. The proposed method achieved high accuracy for the phantom verification. For the clinical cases, the intra- and inter-fractional variations of the fiducial marker were  ⩽3 mm and  ±3 mm in the SI, AP, and LR directions. We proposed a method for the verification of respiratory-gated radiotherapy with SyncTraX using cine EPID images and a log file and showed that this treatment is performed with high accuracy in clinical cases.

  12. Frame average optimization of cine-mode EPID images used for routine clinical in vivo patient dose verification of VMAT deliveries

    Energy Technology Data Exchange (ETDEWEB)

    McCowan, P. M., E-mail: pmccowan@cancercare.mb.ca [Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2, Canada and Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada); McCurdy, B. M. C. [Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2 (Canada); Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada); Department of Radiology, University of Manitoba, 820 Sherbrook Street, Winnipeg, Manitoba R3A 1R9 (Canada)

    2016-01-15

    Purpose: The in vivo 3D dose delivered to a patient during volumetric modulated arc therapy (VMAT) delivery can be calculated using electronic portal imaging device (EPID) images. These images must be acquired in cine-mode (i.e., “movie” mode) in order to capture the time-dependent delivery information. The angle subtended by each cine-mode EPID image during an arc can be changed via the frame averaging number selected within the image acquisition software. A large frame average number will decrease the EPID’s angular resolution and will result in a decrease in the accuracy of the dose information contained within each image. Alternatively, less EPID images acquired per delivery will decrease the overall 3D patient dose calculation time, which is appealing for large-scale clinical implementation. Therefore, the purpose of this study was to determine the optimal frame average value per EPID image, defined as the highest frame averaging that can be used without an appreciable loss in 3D dose reconstruction accuracy for VMAT treatments. Methods: Six different VMAT plans and six different SBRT-VMAT plans were delivered to an anthropomorphic phantom. Delivery was carried out on a Varian 2300ix model linear accelerator (Linac) equipped with an aS1000 EPID running at a frame acquisition rate of 7.5 Hz. An additional PC was set up at the Linac console area, equipped with specialized frame-grabber hardware and software packages allowing continuous acquisition of all EPID frames during delivery. Frames were averaged into “frame-averaged” EPID images using MATLAB. Each frame-averaged data set was used to calculate the in vivo dose to the patient and then compared to the single EPID frame in vivo dose calculation (the single frame calculation represents the highest possible angular resolution per EPID image). A mean percentage dose difference of low dose (<20% prescription dose) and high dose regions (>80% prescription dose) was calculated for each frame averaged

  13. Frame average optimization of cine-mode EPID images used for routine clinical in vivo patient dose verification of VMAT deliveries

    International Nuclear Information System (INIS)

    McCowan, P. M.; McCurdy, B. M. C.

    2016-01-01

    Purpose: The in vivo 3D dose delivered to a patient during volumetric modulated arc therapy (VMAT) delivery can be calculated using electronic portal imaging device (EPID) images. These images must be acquired in cine-mode (i.e., “movie” mode) in order to capture the time-dependent delivery information. The angle subtended by each cine-mode EPID image during an arc can be changed via the frame averaging number selected within the image acquisition software. A large frame average number will decrease the EPID’s angular resolution and will result in a decrease in the accuracy of the dose information contained within each image. Alternatively, less EPID images acquired per delivery will decrease the overall 3D patient dose calculation time, which is appealing for large-scale clinical implementation. Therefore, the purpose of this study was to determine the optimal frame average value per EPID image, defined as the highest frame averaging that can be used without an appreciable loss in 3D dose reconstruction accuracy for VMAT treatments. Methods: Six different VMAT plans and six different SBRT-VMAT plans were delivered to an anthropomorphic phantom. Delivery was carried out on a Varian 2300ix model linear accelerator (Linac) equipped with an aS1000 EPID running at a frame acquisition rate of 7.5 Hz. An additional PC was set up at the Linac console area, equipped with specialized frame-grabber hardware and software packages allowing continuous acquisition of all EPID frames during delivery. Frames were averaged into “frame-averaged” EPID images using MATLAB. Each frame-averaged data set was used to calculate the in vivo dose to the patient and then compared to the single EPID frame in vivo dose calculation (the single frame calculation represents the highest possible angular resolution per EPID image). A mean percentage dose difference of low dose ( 80% prescription dose) was calculated for each frame averaged scenario for each plan. The authors defined their

  14. Epid cine acquisition mode for in vivo dosimetry in dynamic arc radiation therapy

    International Nuclear Information System (INIS)

    Fidanzio, Andrea; Mameli, Alessandra; Placidi, Elisa; Greco, Francesca; Stimato, Gerardina; Gaudino, Diego; Ramella, Sara; D'Angelillo, Rolando; Cellini, Francesco; Trodella, Lucio; Cilla, Savino; Grimaldi, Luca; D'Onofrio, Guido; Azario, Luigi; Piermattei, Angelo

    2008-01-01

    In this paper the cine acquisition mode of an electronic portal imaging device (EPID) has been calibrated and tested to determine the in vivo dose for dynamic conformal arc radiation therapy (DCAT). The EPID cine acquisition mode, that allows a frame acquisition rate of one image every 1.66 s, was studied with a monitor unit rate equal to 100 UM/min. In these conditions good signal stability, ±1% (2SD) evaluated during three months, signal reproducibility within ±0.8% (2SD) and linearity with dose and dose rate within ±1% (2SD) were obtained. The transit signal, S t , (due to the transmitted beam below the phantom) measured by the EPID cine acquisition mode was used to determine, (i) a set of correlation functions, F(w,L), defined as the ratio between S t and the dose at half thickness, D m , measured in solid water phantoms of different thicknesses, w and with square fields of side L, (ii) a set of factors, f(d,L), that take into account the different X-ray scatter contribution from the phantom to the S t signal as a function of the variation, d, of the air gap between the phantom and the EPID. The reconstruction of the isocenter dose, D iso , for DCAT was obtained convolving the transit signal values, obtained at different gantry angles, with the respective reconstruction factors determined by a house-made software. The method was tested with cylindrical and anthropomorphic phantoms and the results show that the reconstructed D iso values can be obtained with an accuracy within ±2.5% in cylindrical phantom and within ±3.4% for anthropomorphic phantom. In conclusion, the transit dosimetry by EPID was assessed to be adequate to perform DCAT in vivo dosimetry, that is not realizable with the other traditional techniques. Moreover, the method proposed here could be implemented to supply in vivo dose values in real time

  15. SU-F-T-263: Dosimetric Characteristics of the Cine Acquisition Mode of An A-Si EPID

    Energy Technology Data Exchange (ETDEWEB)

    Bawazeer, O; Deb, P [RMIT University, Melbourne, VIC (Australia); Sarasanandarajah, S [Peter MacCallum Cancer Institute, Melbourne, Victoria (Australia); Herath, S; Kron, T [Peter MacCallum Cancer Institute, Melbourne, VIC (Australia)

    2016-06-15

    Purpose: To investigate the dosimetric characteristics of Varian a-Si-500 electronic portal imaging device (EPID) operated in cine mode particularly considering linearity with delivered dose, dose rate, field size, phantom thickness, MLC speed and common IMRT fields. Methods: The EPID that attached to a Varian Clinac 21iX linear accelerator, was irradiated with 6 and 18 MV using 600 MU/min. Image acquisition is controlled by the IAS3 software, Trigger delay was 6 ms, BeamOnDelay and FrameStartDelay were zero. Different frame rates were utilized. Cine mode response was calculated using MATLAB as summation of mean pixel values in a region of interest of the acquired images. The performance of cine mode was compared to integrated mode and dose measurements in water using CC13 ionization chamber. Results: Figure1 illustrates that cine mode has nonlinear response for small MU, when delivering 10 MU was about 0.5 and 0.64 for 6 and 18 MV respectively. This is because the missing acquired images that were calculated around four images missing in each delivery. With the increase MU the response became linear and comparable with integrated mode and ionization chamber within 2%. Figure 2 shows that cine mode has comparable response with integrated mode and ionization chamber within 2% with changing dose rate for 10 MU delivered. This indicates that the dose rate change has no effect on nonlinearity of cine mode response. Except nonlinearity, cine mode is well matched to integrated mode response within 2% for field size, phantom thickness, MLC speed dependences. Conclusion: Cine mode has similar dosimetric characteristics to integrated mode with open and IMRT fields, and the main limitation with cine mode is missing images. Therefore, the calibration of EPID images with this mode should be run with large MU, and when IMRT verification field has low MU, the correction for missing images are required.

  16. SU-F-T-263: Dosimetric Characteristics of the Cine Acquisition Mode of An A-Si EPID

    International Nuclear Information System (INIS)

    Bawazeer, O; Deb, P; Sarasanandarajah, S; Herath, S; Kron, T

    2016-01-01

    Purpose: To investigate the dosimetric characteristics of Varian a-Si-500 electronic portal imaging device (EPID) operated in cine mode particularly considering linearity with delivered dose, dose rate, field size, phantom thickness, MLC speed and common IMRT fields. Methods: The EPID that attached to a Varian Clinac 21iX linear accelerator, was irradiated with 6 and 18 MV using 600 MU/min. Image acquisition is controlled by the IAS3 software, Trigger delay was 6 ms, BeamOnDelay and FrameStartDelay were zero. Different frame rates were utilized. Cine mode response was calculated using MATLAB as summation of mean pixel values in a region of interest of the acquired images. The performance of cine mode was compared to integrated mode and dose measurements in water using CC13 ionization chamber. Results: Figure1 illustrates that cine mode has nonlinear response for small MU, when delivering 10 MU was about 0.5 and 0.64 for 6 and 18 MV respectively. This is because the missing acquired images that were calculated around four images missing in each delivery. With the increase MU the response became linear and comparable with integrated mode and ionization chamber within 2%. Figure 2 shows that cine mode has comparable response with integrated mode and ionization chamber within 2% with changing dose rate for 10 MU delivered. This indicates that the dose rate change has no effect on nonlinearity of cine mode response. Except nonlinearity, cine mode is well matched to integrated mode response within 2% for field size, phantom thickness, MLC speed dependences. Conclusion: Cine mode has similar dosimetric characteristics to integrated mode with open and IMRT fields, and the main limitation with cine mode is missing images. Therefore, the calibration of EPID images with this mode should be run with large MU, and when IMRT verification field has low MU, the correction for missing images are required.

  17. Cine EPID evaluation of two non-commercial techniques for DIBH

    Energy Technology Data Exchange (ETDEWEB)

    Jensen, Christopher; Urribarri, Jaime; Cail, Daniel; Rottmann, Joerg; Mishra, Pankaj; Lingos, Tatiana; Niedermayr, Thomas; Berbeco, Ross, E-mail: rberbeco@lroc.harvard.edu [Brigham and Women' s Hospital, Dana-Farber Cancer Institute, Harvard Medical School, Boston, Massachusetts 02115 (United States)

    2014-02-15

    Purpose: To evaluate the efficacy of two noncommercial techniques for deep inspiration breathhold (DIBH) treatment of left-sided breast cancer (LSBC) usingcine electronic portal imaging device (EPID) images. Methods: 23 875 EPID images of 65 patients treated for LSBC at two different cancer treatment centers were retrieved. At the Milford Regional Cancer Center, DIBH stability was maintained by visual alignment of inroom lasers and patient skin tattoos (TAT). At the South Shore Hospital, a distance-measuring laser device (RTSSD) was implemented. For both centers,cine EPID images were acquired at least once per week during beam-on. Chest wall position relative to image boundary was measured and tracked over the course of treatment for every patient and treatment fraction for which data were acquired. Results: Median intrabeam chest motion was 0.31 mm for the TAT method and 0.37 mm for the RTSSD method. The maximum excursions exceeded our treatment protocol threshold of 3 mm in 0.3% of cases (TAT) and 1.2% of cases (RTSSD). The authors did not observe a clinically significant difference between the two datasets. Conclusions: Both noncommercial techniques for monitoring the DIBH location provided DIBH stability within the predetermined treatment protocol parameters (<3 mm). The intreatment imaging offered by the EPID operating incine mode facilitates retrospective analysis and validation of both techniques.

  18. Cine EPID evaluation of two non-commercial techniques for DIBH

    International Nuclear Information System (INIS)

    Jensen, Christopher; Urribarri, Jaime; Cail, Daniel; Rottmann, Joerg; Mishra, Pankaj; Lingos, Tatiana; Niedermayr, Thomas; Berbeco, Ross

    2014-01-01

    Purpose: To evaluate the efficacy of two noncommercial techniques for deep inspiration breathhold (DIBH) treatment of left-sided breast cancer (LSBC) usingcine electronic portal imaging device (EPID) images. Methods: 23 875 EPID images of 65 patients treated for LSBC at two different cancer treatment centers were retrieved. At the Milford Regional Cancer Center, DIBH stability was maintained by visual alignment of inroom lasers and patient skin tattoos (TAT). At the South Shore Hospital, a distance-measuring laser device (RTSSD) was implemented. For both centers,cine EPID images were acquired at least once per week during beam-on. Chest wall position relative to image boundary was measured and tracked over the course of treatment for every patient and treatment fraction for which data were acquired. Results: Median intrabeam chest motion was 0.31 mm for the TAT method and 0.37 mm for the RTSSD method. The maximum excursions exceeded our treatment protocol threshold of 3 mm in 0.3% of cases (TAT) and 1.2% of cases (RTSSD). The authors did not observe a clinically significant difference between the two datasets. Conclusions: Both noncommercial techniques for monitoring the DIBH location provided DIBH stability within the predetermined treatment protocol parameters (<3 mm). The intreatment imaging offered by the EPID operating incine mode facilitates retrospective analysis and validation of both techniques

  19. Digital cine-imaging

    International Nuclear Information System (INIS)

    Masuda, Kazuhiro

    1992-01-01

    Digitization of fluoroscopic images has been developed for the digital cine imaging system as a result of the computer technology, television technology, and popularization of interventional radiology. Present digital cine imaging system is able to offer images similar to cine film because of the higher operatability and better image quality with the development of interventional radiology. As a result, its higher usefulness for catheter diagnosis examination except for interventional radiology was reported, and the possibility of having filmless cine is close to becoming a reality. However several problems have been pointed out, such as spatial resolution, time resolution, storage and exchangeability of data, disconsolidated viewing functions, etc. Anyhow, digital cine imaging system has some unresolved points and lots the needs to be discussed. The tendency of digitization is the passage of the time and we have to promote a study for more useful digital cine imaging system in team medical treatment which centers on the patients. (author)

  20. Craniocaudal Safety Margin Calculation Based on Interfractional Changes in Tumor Motion in Lung SBRT Assessed With an EPID in Cine Mode

    International Nuclear Information System (INIS)

    Ueda, Yoshihiro; Miyazaki, Masayoshi; Nishiyama, Kinji; Suzuki, Osamu; Tsujii, Katsutomo; Miyagi, Ken

    2012-01-01

    Purpose: To evaluate setup error and interfractional changes in tumor motion magnitude using an electric portal imaging device in cine mode (EPID cine) during the course of stereotactic body radiation therapy (SBRT) for non–small-cell lung cancer (NSCLC) and to calculate margins to compensate for these variations. Materials and Methods: Subjects were 28 patients with Stage I NSCLC who underwent SBRT. Respiratory-correlated four-dimensional computed tomography (4D-CT) at simulation was binned into 10 respiratory phases, which provided average intensity projection CT data sets (AIP). On 4D-CT, peak-to-peak motion of the tumor (M-4DCT) in the craniocaudal direction was assessed and the tumor center (mean tumor position [MTP]) of the AIP (MTP-4DCT) was determined. At treatment, the tumor on cone beam CT was registered to that on AIP for patient setup. During three sessions of irradiation, peak-to-peak motion of the tumor (M-cine) and the mean tumor position (MTP-cine) were obtained using EPID cine and in-house software. Based on changes in tumor motion magnitude (∆M) and patient setup error (∆MTP), defined as differences between M-4DCT and M-cine and between MTP-4DCT and MTP-cine, a margin to compensate for these variations was calculated with Stroom’s formula. Results: The means (±standard deviation: SD) of M-4DCT and M-cine were 3.1 (±3.4) and 4.0 (±3.6) mm, respectively. The means (±SD) of ∆M and ∆MTP were 0.9 (±1.3) and 0.2 (±2.4) mm, respectively. Internal target volume-planning target volume (ITV-PTV) margins to compensate for ∆M, ∆MTP, and both combined were 3.7, 5.2, and 6.4 mm, respectively. Conclusion: EPID cine is a useful modality for assessing interfractional variations of tumor motion. The ITV-PTV margins to compensate for these variations can be calculated.

  1. Evaluation of motion measurement using cine MRI for image guided stereotactic body radiotherapy on a new phantom platform

    Science.gov (United States)

    Cai, Jing; Wang, Ziheng; Yin, Fang-Fang

    2011-01-01

    The objective of this study is to investigate accuracy of motion tracking of cine magnetic resonance imaging (MRI) for image-guided stereotactic body radiotherapy. A phantom platform was developed in this work to fulfill the goal. The motion phantom consisted of a platform, a solid thread, a motor and a control system that can simulate motion in various modes. To validate its reproducibility, the phantom platform was setup three times and imaged with fluoroscopy using an electronic portal imaging device (EPID) for the same motion profile. After the validation test, the phantom platform was evaluated using cine MRI at 2.5 frames/second on a 1.5T GE scanner using five different artificial profiles and five patient profiles. The above profiles were again measured with EPID fluoroscopy and used as references. Discrepancies between measured profiles from cine MRI and EPID were quantified using root-mean-square (RMS) and standard deviation (SD). Pearson’s product moment correlational analysis was used to test correlation. The standard deviation for the reproducibility test was 0.28 mm. The discrepancies (RMS) between all profiles measured by cine MRI and EPID fluoroscopy ranged from 0.30 to 0.49 mm for artificial profiles and ranged from 0.75 to 0.91 mm for five patient profiles. The cine MRI sequence could precisely track phantom motion and the proposed motion phantom was feasible to evaluate cine MRI accuracy. PMID:29296304

  2. Sci-Fri PM: Radiation Therapy, Planning, Imaging, and Special Techniques - 11: Quantification of chest wall motion during deep inspiration breast hold treatments using cine EPID images and a physics based algorithm

    Energy Technology Data Exchange (ETDEWEB)

    Alpuche Aviles, Jorge E.; VanBeek, Timothy [CancerCare Manitoba, Winnipeg (Canada); Sasaki, David; Rivest, Ryan; Akra, Mohamed [CancerCare Manitoba, Winnipeg (Canada); University of Manitoba, Winnipeg (Canada)

    2016-08-15

    Purpose: This work presents an algorithm used to quantify intra-fraction motion for patients treated using deep inspiration breath hold (DIBH). The algorithm quantifies the position of the chest wall in breast tangent fields using electronic portal images. Methods: The algorithm assumes that image profiles, taken along a direction perpendicular to the medial border of the field, follow a monotonically and smooth decreasing function. This assumption is invalid in the presence of lung and can be used to calculate chest wall position. The algorithm was validated by determining the position of the chest wall for varying field edge positions in portal images of a thoracic phantom. The algorithm was used to quantify intra-fraction motion in cine images for 7 patients treated with DIBH. Results: Phantom results show that changes in the distance between chest wall and field edge were accurate within 0.1 mm on average. For a fixed field edge, the algorithm calculates the position of the chest wall with a 0.2 mm standard deviation. Intra-fraction motion for DIBH patients was within 1 mm 91.4% of the time and within 1.5 mm 97.9% of the time. The maximum intra-fraction motion was 3.0 mm. Conclusions: A physics based algorithm was developed and can be used to quantify the position of chest wall irradiated in tangent portal images with an accuracy of 0.1 mm and precision of 0.6 mm. Intra-fraction motion for patients treated with DIBH at our clinic is less than 3 mm.

  3. Cine MR imaging in valvular heart disease

    Energy Technology Data Exchange (ETDEWEB)

    Nishimura, Tsunehiko; Yamada, Naoaki; Itoh, Akira; Miyatake, Kunio

    1989-01-01

    Cine MR Imaging was carried out using FLASH (fast low angle shot) which employes TE of 16 msec and TR of 30/similar to/40 msec. Regurgitant jet was visible as discrete area of low signal intensity extending from the incompetent valve into the respective cardiac chamber. In 20 patients with mitral regurgitation, the correlation of the length and area of mitral jet by cine MR and color doppler mapping was 0.74 and 0.71, respectively. Cine MR imaging is a promising modality for detection and quantification of valvular heart disease.

  4. Cine MR imaging of valvular heart disease

    International Nuclear Information System (INIS)

    Underwood, S.R.; Firmin, D.N.; Mohiaddin, R.H.; Klipstein, R.H.; Bogren, H.G.; Rees, R.S.O.; Longmore, D.B.

    1987-01-01

    Cine MR imaging shows details of cardiac valve motion and the turbulence caused by regurgitation and stenosis. In 30 patients, the area of turbulence proximal to regurgitant valves correlated well with the regurgitant fraction measured from right and left ventricular stroke volumes. Turbulence distal to stenotic valves did not correlate with the pressure gradient because abnormal valves that were not stenosed produced turbulence. Valve motion and the presence of thrombus were easier to assess than in conventional images. Cine MR imaging provides an important addition to conventional imaging in valvular regurgitation, but pressure gradients cannot be assessed in valvular stenosis

  5. Assessment of voluntary deep inspiration breath-hold with CINE imaging for breast radiotherapy.

    Science.gov (United States)

    Estoesta, Reuben Patrick; Attwood, Lani; Naehrig, Diana; Claridge-Mackonis, Elizabeth; Odgers, David; Martin, Darren; Pham, Melissa; Toohey, Joanne; Carroll, Susan

    2017-10-01

    Deep Inspiration Breath-Hold (DIBH) techniques for breast cancer radiation therapy (RT) have reduced cardiac dose compared to Free Breathing (FB). Recently, a voluntary deep inspiration breath-hold (vDIBH) technique was established using in-room lasers and skin tattoos to monitor breath-hold. An in-house quality assessment of positional reproducibility during RT delivery with vDIBH in patients with left-sided breast cancer was evaluated. The electronic portal imaging device (EPID) was used in cinematographic (CINE) mode to capture a sequence of images during beam delivery. Weekly CINE images were retrospectively assessed for 20 left-sided breast cancer patients receiving RT in vDIBH, and compared with CINE images of 20 patients treated in FB. The intra-beam motion was assessed and the distance from the beam central axis (CA) to the internal chest wall (ICW) was measured on each CINE image. These were then compared to the planned distance on digitally reconstructed radiograph (DRR). The maximum intra-beam motion for any one patient measurement was 0.30 cm for vDIBH and 0.20 cm for FB. The mean difference between the distance from the CA to ICW on DRR and the equivalent distance on CINE imaging (as treated) was 0.28 cm (SD 0.17) for vDIBH patients and 0.25 cm (SD 0.14) for FB patients (P = 0.458). The measured values were comparable for patients undergoing RT in vDIBH, and for those in FB. This quality assessment showed that using in-room lasers and skin tattoos to independently monitor breath-hold in vDIBH as detected by 'on-treatment' CINE imaging is safe and effective. © 2017 The Royal Australian and New Zealand College of Radiologists.

  6. Cine MR imaging in mitral valve prolapse

    International Nuclear Information System (INIS)

    Kumai, Toshihiko

    1993-01-01

    This study was undertaken to assess the ability of cine MR imaging to evaluate the direction, timing, and severity of mitral regurgitation in patients with mitral valve prolapse (MVP). The population of this study was 33 patients with MVP diagnosed by two-dimensional echocardiography and 10 patients with rheumatic mitral valve disease (MSR) for comparison. 7 patients with MVP and 5 with MSR had atrial fibrillation and/or history of congestive heart failure as complications. Mitral regurgitation was graded for severity by color Doppler flow imaging in all patients. Direction and size of systolic flow void in the left atrium were analyzed by contiguous multilevel cine MR images and the maximum volumes of flow void and left atrium were measured. Although flow void was found at the center of the left atrium in most of MSR, it was often directed along the postero-caudal atrial wall in anterior leaflet prolapse and along the anterocranial atrial wall in posterior leaflet prolapse. In MVP, the maximum volume of flow void was often seen in late systole. The maximum volume of flow void and that of left atrium were significantly larger in patients with atrial fibrillation and/or history of congestive heart failure. The length and volume of flow void were increased with clinical severity and degree of regurgitation determined by color Doppler flow imaging. Thus cine MR imaging provides a useful means for detection and semiquantitative evaluation of mitral regurgitation in subjects with MVP. (author)

  7. Cine MR imaging of internal derangements of the knee

    International Nuclear Information System (INIS)

    Niitsu, M.; Anno, I.; Ishikawa, N.; Akisada, M.; Fukubayashi, T.

    1990-01-01

    This paper assesses internal derangements of the knee joint by a new method of cine MR imaging. Cine MR imaging, involving cine acquisition of kinematic sagittal images during knee movement, was used to evaluate 51 symptomatic patients and 19 healthy subjects. Cine MR images of cruciate ligament fears showed disappearance of low-intensity bundles during knee movement, periodic appearance of joint fluid in the area of the ligament, and irregular tibial movement. Cine MR images of meniscal tears showed independent movement of meniscal fragments and periodic tear opening and closing. Twenty-seven of the 28 anterior cruciate ligament (ACL) tears confirmed arthroscopically were correctly identified with cine MR imaging, with a sensitivity of 96% and a specificity of 92%

  8. Cine MR imaging of the cardiovascular system

    International Nuclear Information System (INIS)

    Nishimura, Fumiaki; Yoshino, Yasushi; Mihara, Junji; Ichikawa, Seiichi; Kimura, Masahiko; Yano, Masao; Umeda, Masahiro; Oouchi, Toshihiro

    1990-01-01

    In recent years cine magnetic resonance imaging (MRI) has developed as a high-speed imaging technique that provides a high intensity signal even at a short repetition time (20-30 msec) by using an excited pulse with a small flip angle according to the gradient echo method, enabling about 20 to 30 continuous images of the same section per one cardiac cycle to be taken. On cine display of these continuous images, information concerning blood flow shown by a high intensity signal in comparison with that of the myocardium and vascular wall is obtained with high temporal resolution along with anatomical information. The present study reports the clinical usefulness of cine MRI in today's situation, inculding the following: calculation of the left ventricular ejection fraction and pulmonary-to-systemic flow ratio in congenital shunt disease by integration of the area of multisections through application of Simpson's method; diagnosis of the severity of valvular regurgitation, evaluation of stenosal diseases, and diagnosis of inflow from the fissured entry of dissecting aortic aneurysm by evaluating of an area of low intensity signal, probably based on the high velocity or turbulent blood flow: and evaluation of patency of the internal mammary artery bypass graft of the basis of the possible visualization of even thin blood vessels because of the high intensity signal of blood flow. In particular, the characteristics of this procedures are described by comparing it with other technologies in the field of diseases of valvular regurgitation. (author)

  9. Cine MR imaging of cerebrospinal fluid flow

    International Nuclear Information System (INIS)

    Kawahara, K.; Yoshikawa, A.; Maeda, M.; Hamachi, J.; Morita, R.; Mishima, T.; Yamada, R.

    1988-01-01

    Cine MR imaging was used to study cerebrospinal fluid (CSF) flow in five patients with various kinds of hydrocephalus, 14 patients with brain atrophy, and 26 healthy subjects. For each study, sagittal and axial sections were obtained. In this study, the authors detected cerebrospinal fluid flow, apparent as high signal intensity, and its direction. They detected the outflow from the foramen of Monro, aqueduct, and foramen of Magendie in sagittal cine images of patients with those with normal-pressure hydrocehalus (NPH), those with brain atrophy, and controls. Especially in patients with NPH, the markedly high signal intensity from CSF flow was observeed in the third and fourth ventricles. In patients with obstructive hydrocephalus, the fourth ventricle tumor, the outflow of CSF was not present. In axial sections, the signal intensity of aqueduct in patients with obstructive hydrocephalus was lower, whereas that in patients with NPH, controls, and those with atrophy was higher than in the midbrain. Cine MR imaging is a very attractive technique in diagnosing hydrocephalus and other abnormal conditions

  10. Characterization of a novel EPID designed for simultaneous imaging and dose verification in radiotherapy

    International Nuclear Information System (INIS)

    Blake, Samuel J.; McNamara, Aimee L.; Deshpande, Shrikant; Holloway, Lois; Greer, Peter B.; Kuncic, Zdenka; Vial, Philip

    2013-01-01

    Purpose: Standard amorphous silicon electronic portal imaging devices (a-Si EPIDs) are x-ray imagers used frequently in radiotherapy that indirectly detect incident x-rays using a metal plate and phosphor screen. These detectors may also be used as two-dimensional dosimeters; however, they have a well-characterized nonwater-equivalent dosimetric response. Plastic scintillating (PS) fibers, on the other hand, have been shown to respond in a water-equivalent manner to x-rays in the energy range typically encountered during radiotherapy. In this study, the authors report on the first experimental measurements taken with a novel prototype PS a-Si EPID developed for the purpose of performing simultaneous imaging and dosimetry in radiotherapy. This prototype employs an array of PS fibers in place of the standard metal plate and phosphor screen. The imaging performance and dosimetric response of the prototype EPID were evaluated experimentally and compared to that of the standard EPID.Methods: Clinical 6 MV photon beams were used to first measure the detector sensitivity, linearity of dose response, and pixel noise characteristics of the prototype and standard EPIDs. Second, the dosimetric response of each EPID was evaluated relative to a reference water-equivalent dosimeter by measuring the off-axis and field size response in a nontransit configuration, along with the off-axis, field size, and transmission response in a transit configuration using solid water blocks. Finally, the imaging performance of the prototype and standard EPIDs was evaluated quantitatively by using an image quality phantom to measure the contrast to noise ratio (CNR) and spatial resolution of images acquired with each detector, and qualitatively by using an anthropomorphic phantom to acquire images representative of human anatomy.Results: The prototype EPID's sensitivity was 0.37 times that of the standard EPID. Both EPIDs exhibited responses that were linear with delivered dose over a range of 1

  11. Automatic segmentation and disease classification using cardiac cine MR images

    NARCIS (Netherlands)

    Wolterink, Jelmer M.; Leiner, Tim; Viergever, Max A.; Išgum, Ivana

    2018-01-01

    Segmentation of the heart in cardiac cine MR is clinically used to quantify cardiac function. We propose a fully automatic method for segmentation and disease classification using cardiac cine MR images. A convolutional neural network (CNN) was designed to simultaneously segment the left ventricle

  12. SU-C-202-07: Protocol and Hardware for Improved Flood Field Calibration of TrueBeam FFF Cine Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Adamson, J; Faught, A; Yin, F [Duke University Medical Center, Durham, NC (United States)

    2016-06-15

    Purpose: Flattening filter free photon energies are commonly used for high dose treatments such as SBRT, where localization accuracy is essential. Often, MV cine imaging may be employed to verify correct localization. TrueBeam Electronic Portal Imaging Devices (EPIDs) equipped with the 40×30cm{sup 2} Image Detection Unit (IDU) are prone to image saturation at the image center especially for higher dose rates. While saturation often does not occur for cine imaging during treatment because the beam is attenuated by the patient, the flood field calibration is affected when the standard calibration procedure is followed. Here we describe the hardware and protocol to achieve improved image quality for this model of TrueBeam EPID. Methods: A stainless steel filter of uniform thickness was designed to have sufficient attenuation to avoid panel saturation for both 6XFFF and 10XFFF at the maximum dose rates (1400 MU/min & 2400 MU/min, respectively). The cine imaging flood field calibration was then acquired with the filter in place for the FFF energies under the standard calibration geometry (SDD=150cm). Image quality during MV cine was assessed with & without the modified flood field calibration using a low contrast resolution phantom and an anthropomorphic phantom. Results: When the flood field is acquired using the standard procedure (no filter in place), a pixel gain artifact is clearly present in the image center (r=3cm for 10XFFF at 2400 MU/min) which appears similar to and may be mis-attributed to panel saturation in the subject image. The artifact obscured all low contrast inserts at the image center and was also visible on the anthropomorphic phantom. Using the filter for flood field calibration eliminated the artifact. Conclusion: Use of a modified flood field calibration procedure improves image quality for cine MV imaging with TrueBeams equipped with the 40×30cm{sup 2} IDU.

  13. Quality control of portal imaging with PTW EPID QC PHANTOM registered

    International Nuclear Information System (INIS)

    Pesznyak, Csilla; Kiraly, Reka; Polgar, Istvan; Zarand, Pal; Mayer, Arpad; Fekete, Gabor; Mozes, Arpad; Kiss, Balazs

    2009-01-01

    Purpose: quality assurance (QA) and quality control (QC) of different electronic portal imaging devices (EPID) and portal images with the PTW EPID QC PHANTOM registered . Material and methods: characteristic properties of images of different file formats were measured on Siemens OptiVue500aSi registered , Siemens BeamView Plus registered , Elekta iView registered , and Varian PortalVision trademark and analyzed with the epidSoft registered 2.0 program in four radiation therapy centers. The portal images were taken with Kodak X-OMAT V registered and the Kodak Portal Localisation ReadyPack registered films and evaluated with the same program. Results: the optimal exposition both for EPIDs and portal films of different kind was determined. For double exposition, the 2+1 MU values can be recommended in the case of Siemens OptiVue500aSi registered , Elekta iView registered and Kodak Portal Localisation ReadyPack registered films, while for Siemens BeamView Plus registered , Varian PortalVision trademark and Kodak X-OMAT V registered film 7+7 MU is recommended. Conclusion: the PTW EPID QC PHANTOM registered can be used not only for amorphous silicon EPIDs but also for images taken with a video-based system or by using an ionization chamber matrix or for portal film. For analysis of QC tests, a standardized format (used at the acceptance test) should be applied, as the results are dependent on the file format used. (orig.)

  14. Development of a daily dosimetric control for radiation therapy using an electronic portal imaging device (EPID)

    International Nuclear Information System (INIS)

    Saboori, Mohammadsaeed

    2015-01-01

    Electronic Portal Imaging Devices (EPIDs) can be used to perform dose measurements during radiation therapy treatments if dedicated calibration and correction procedures are applied. The purpose of this study was to provide a new calibration and correction model for an amorphous silicon (a-Si) EPID for use in transit dose verification of step-and-shoot intensity modulated radiation therapy (IMRT). A model was created in a commercial treatment planning system to calculate the nominal two-dimensional (2D) dose map of each radiation field at the EPID level. The EPID system was calibrated and correction factors were determined using a reference set-up, which consisted a patient phantom and an EPID phantom. The advantage of this method is that for the calibration, the actual beam spectrum is used to mimic a patient measurement. As proof-of-principle, the method was tested for the verification of two 7-field IMRT treatment plans with tumor sites in the head-and-neck and pelvic region. Predicted and measured EPID responses were successfully compared to the nominal data from treatment planning using dose difference maps and gamma analyses. Based on our result it can be concluded that this new method of 2D EPID dosimetry is a potential tool for simple patient treatment fraction dose verification.

  15. Four-dimensional dose reconstruction through in vivo phase matching of cine images of electronic portal imaging device.

    Science.gov (United States)

    Yoon, Jihyung; Jung, Jae Won; Kim, Jong Oh; Yi, Byong Yong; Yeo, Inhwan

    2016-07-01

    A method is proposed to reconstruct a four-dimensional (4D) dose distribution using phase matching of measured cine images to precalculated images of electronic portal imaging device (EPID). (1) A phantom, designed to simulate a tumor in lung (a polystyrene block with a 3 cm diameter embedded in cork), was placed on a sinusoidally moving platform with an amplitude of 1 cm and a period of 4 s. Ten-phase 4D computed tomography (CT) images of the phantom were acquired. A planning target volume (PTV) was created by adding a margin of 1 cm around the internal target volume of the tumor. (2) Three beams were designed, which included a static beam, a theoretical dynamic beam, and a planning-optimized dynamic beam (PODB). While the theoretical beam was made by manually programming a simplistic sliding leaf motion, the planning-optimized beam was obtained from treatment planning. From the three beams, three-dimensional (3D) doses on the phantom were calculated; 4D dose was calculated by means of the ten phase images (integrated over phases afterward); serving as "reference" images, phase-specific EPID dose images under the lung phantom were also calculated for each of the ten phases. (3) Cine EPID images were acquired while the beams were irradiated to the moving phantom. (4) Each cine image was phase-matched to a phase-specific CT image at which common irradiation occurred by intercomparing the cine image with the reference images. (5) Each cine image was used to reconstruct dose in the phase-matched CT image, and the reconstructed doses were summed over all phases. (6) The summation was compared with forwardly calculated 4D and 3D dose distributions. Accounting for realistic situations, intratreatment breathing irregularity was simulated by assuming an amplitude of 0.5 cm for the phantom during a portion of breathing trace in which the phase matching could not be performed. Intertreatment breathing irregularity between the time of treatment and the time of planning CT was

  16. Four-dimensional dose reconstruction through in vivo phase matching of cine images of electronic portal imaging device

    Energy Technology Data Exchange (ETDEWEB)

    Yoon, Jihyung; Jung, Jae Won, E-mail: jungj@ecu.edu [Department of Physics, East Carolina University, Greenville, North Carolina 27858 (United States); Kim, Jong Oh [Department of Radiation Oncology, University of Pittsburgh Cancer Institute, Pittsburgh, Pennsylvania 15232 (United States); Yi, Byong Yong [Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland 21201 (United States); Yeo, Inhwan [Department of Radiation Medicine, Loma Linda University Medical Center, Loma Linda, California 92354 (United States)

    2016-07-15

    Purpose: A method is proposed to reconstruct a four-dimensional (4D) dose distribution using phase matching of measured cine images to precalculated images of electronic portal imaging device (EPID). Methods: (1) A phantom, designed to simulate a tumor in lung (a polystyrene block with a 3 cm diameter embedded in cork), was placed on a sinusoidally moving platform with an amplitude of 1 cm and a period of 4 s. Ten-phase 4D computed tomography (CT) images of the phantom were acquired. A planning target volume (PTV) was created by adding a margin of 1 cm around the internal target volume of the tumor. (2) Three beams were designed, which included a static beam, a theoretical dynamic beam, and a planning-optimized dynamic beam (PODB). While the theoretical beam was made by manually programming a simplistic sliding leaf motion, the planning-optimized beam was obtained from treatment planning. From the three beams, three-dimensional (3D) doses on the phantom were calculated; 4D dose was calculated by means of the ten phase images (integrated over phases afterward); serving as “reference” images, phase-specific EPID dose images under the lung phantom were also calculated for each of the ten phases. (3) Cine EPID images were acquired while the beams were irradiated to the moving phantom. (4) Each cine image was phase-matched to a phase-specific CT image at which common irradiation occurred by intercomparing the cine image with the reference images. (5) Each cine image was used to reconstruct dose in the phase-matched CT image, and the reconstructed doses were summed over all phases. (6) The summation was compared with forwardly calculated 4D and 3D dose distributions. Accounting for realistic situations, intratreatment breathing irregularity was simulated by assuming an amplitude of 0.5 cm for the phantom during a portion of breathing trace in which the phase matching could not be performed. Intertreatment breathing irregularity between the time of treatment and the

  17. Tracking tumor boundary in MV-EPID images without implanted markers: A feasibility study

    International Nuclear Information System (INIS)

    Zhang, Xiaoyong; Homma, Noriyasu; Ichiji, Kei; Takai, Yoshihiro; Yoshizawa, Makoto

    2015-01-01

    Purpose: To develop a markerless tracking algorithm to track the tumor boundary in megavoltage (MV)-electronic portal imaging device (EPID) images for image-guided radiation therapy. Methods: A level set method (LSM)-based algorithm is developed to track tumor boundary in EPID image sequences. Given an EPID image sequence, an initial curve is manually specified in the first frame. Driven by a region-scalable energy fitting function, the initial curve automatically evolves toward the tumor boundary and stops on the desired boundary while the energy function reaches its minimum. For the subsequent frames, the tracking algorithm updates the initial curve by using the tracking result in the previous frame and reuses the LSM to detect the tumor boundary in the subsequent frame so that the tracking processing can be continued without user intervention. The tracking algorithm is tested on three image datasets, including a 4-D phantom EPID image sequence, four digitally deformable phantom image sequences with different noise levels, and four clinical EPID image sequences acquired in lung cancer treatment. The tracking accuracy is evaluated based on two metrics: centroid localization error (CLE) and volume overlap index (VOI) between the tracking result and the ground truth. Results: For the 4-D phantom image sequence, the CLE is 0.23 ± 0.20 mm, and VOI is 95.6% ± 0.2%. For the digital phantom image sequences, the total CLE and VOI are 0.11 ± 0.08 mm and 96.7% ± 0.7%, respectively. In addition, for the clinical EPID image sequences, the proposed algorithm achieves 0.32 ± 0.77 mm in the CLE and 72.1% ± 5.5% in the VOI. These results demonstrate the effectiveness of the authors’ proposed method both in tumor localization and boundary tracking in EPID images. In addition, compared with two existing tracking algorithms, the proposed method achieves a higher accuracy in tumor localization. Conclusions: In this paper, the authors presented a feasibility study of tracking

  18. SU-F-T-261: Reconstruction of Initial Photon Fluence Based On EPID Images

    Energy Technology Data Exchange (ETDEWEB)

    Seliger, T; Engenhart-Cabillic, R [Philipp University of Marburg, Marburg (Germany); Czarnecki, D; Maeder, U; Zink, K [Technische Hochschule Mittelhessen - University of Applied Sciences, Giessen (Germany); Kussaether, R [MedCom GmbH, Darmstadt (Germany); Poppe, B [University Hospital for Medical Radiation Physics, Pius-Hospital, Medical Campus, Carl von Ossietzky University of Oldenburg (Germany)

    2016-06-15

    Purpose: Verifying an algorithm to reconstruct relative initial photon fluence for clinical use. Clinical EPID and CT images were acquired to reconstruct an external photon radiation treatment field. The reconstructed initial photon fluence could be used to verify the treatment or calculate the applied dose to the patient. Methods: The acquired EPID images were corrected for scatter caused by the patient and the EPID with an iterative reconstruction algorithm. The transmitted photon fluence behind the patient was calculated subsequently. Based on the transmitted fluence the initial photon fluence was calculated using a back-projection algorithm which takes the patient geometry and its energy dependent linear attenuation into account. This attenuation was gained from the acquired cone-beam CT or the planning CT by calculating a water-equivalent radiological thickness for each irradiation direction. To verify the algorithm an inhomogeneous phantom consisting of three inhomogeneities was irradiated by a static 6 MV photon field and compared to a reference flood field image. Results: The mean deviation between the reconstructed relative photon fluence for the inhomogeneous phantom and the flood field EPID image was 3% rising up to 7% for off-axis fluence. This was probably caused by the used clinical EPID calibration, which flattens the inhomogeneous fluence profile of the beam. Conclusion: In this clinical experiment the algorithm achieved good results in the center of the field while it showed high deviation of the lateral fluence. This could be reduced by optimizing the EPID calibration, considering the off-axis differential energy response. In further progress this and other aspects of the EPID, eg. field size dependency, CT and dose calibration have to be studied to realize a clinical acceptable accuracy of 2%.

  19. Evaluation of aortic regurgitation using cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Tamai, Takuya; Konishi, Tokuji; Okamoto, Shinya; Sakuma, Hajime; Takeda, Kan; Nakano, Takeshi

    1993-01-01

    Cine magnetic resonance imaging (MRI) was used to assess aortic regurgitation (AR) in 13 patients with valvular disease and 3 normal subjects, and the results were compared to color Doppler flow mapping findings. AR produced a signal void in the left ventricle during the diastolic phase in all patients by MRI. There were no false positive or negative results compared with echocardiographic findings. Visual grading of cine MRI gave results similar to color flow Doppler echocardiography (88%). The distance and the area of aortic regurgitation using MRI correlated well with color Doppler flow mapping (r=0.82 and 0.88). However, measurements of distance and area by color flow Doppler tended to be larger than those by cine MRI. With current techniques echocardiography may overestimate the severity of AR as compared with cine MRI. In addition, MRI gives clinically useful information in patients in whom transthoracic Doppler echocardiography is not adequate. (author)

  20. Technical Note: Measurement of common carotid artery lumen dynamics using black-blood MR cine imaging.

    Science.gov (United States)

    Dai, Erpeng; Dong, Li; Zhang, Zhe; Li, Lyu; Zhang, Hui; Zhao, Xihai; Wang, Jinnan; Yuan, Chun; Guo, Hua

    2017-03-01

    To demonstrate the feasibility of measuring the common carotid artery (CCA) lumen dynamics using a black-blood cine (BB-cine) imaging method. Motion-sensitized driven-equilibrium (MSDE) prepared spoiled gradient sequence was used for the BB-cine imaging. CCAs of eleven healthy volunteers were studied using this method. Lumen dynamics, including lumen area evolution waveforms and distension values, were measured and evaluated by comparing this method with bright-blood cine (BrB-cine) imaging. Compared with the BrB-cine images, flow artifacts were effectively suppressed in the BB-cine images. BrB-cine images generally show larger lumen areas than BB-cine images. The lumen area waveforms and distension measurements from BB-cine imaging showed smaller variances among different subjects than BrB-cine imaging. The proposed BB-cine imaging technique can suppress the flow artifacts effectively and reduce the partial volume effects from the vessel wall. This might allow more accurate lumen dynamics measurements than traditional BrB-cine imaging, which may further be valuable for investigating biomechanical and functional properties of the cardiovascular system. © 2017 American Association of Physicists in Medicine.

  1. Quality control of portal imaging with PTW EPID QC PHANTOM {sup registered}

    Energy Technology Data Exchange (ETDEWEB)

    Pesznyak, Csilla; Kiraly, Reka; Polgar, Istvan; Zarand, Pal; Mayer, Arpad [Inst. of Oncoradiology, Uzsoki Hospital, Budapest (Hungary); Fekete, Gabor [Dept. of Oncotherapy, Univ. of Szeged (Hungary); Mozes, Arpad [Oncology Center, Kalman Pandy County Hospital, Gyula (Hungary); Kiss, Balazs [Dept. of Radiation Oncology, Markusovszky County Hospital, Szombathely (Hungary)

    2009-01-15

    Purpose: quality assurance (QA) and quality control (QC) of different electronic portal imaging devices (EPID) and portal images with the PTW EPID QC PHANTOM {sup registered}. Material and methods: characteristic properties of images of different file formats were measured on Siemens OptiVue500aSi {sup registered}, Siemens BeamView Plus {sup registered}, Elekta iView {sup registered}, and Varian PortalVision trademark and analyzed with the epidSoft {sup registered} 2.0 program in four radiation therapy centers. The portal images were taken with Kodak X-OMAT V {sup registered} and the Kodak Portal Localisation ReadyPack {sup registered} films and evaluated with the same program. Results: the optimal exposition both for EPIDs and portal films of different kind was determined. For double exposition, the 2+1 MU values can be recommended in the case of Siemens OptiVue500aSi {sup registered}, Elekta iView {sup registered} and Kodak Portal Localisation ReadyPack {sup registered} films, while for Siemens BeamView Plus {sup registered}, Varian PortalVision trademark and Kodak X-OMAT V {sup registered} film 7+7 MU is recommended. Conclusion: the PTW EPID QC PHANTOM {sup registered} can be used not only for amorphous silicon EPIDs but also for images taken with a video-based system or by using an ionization chamber matrix or for portal film. For analysis of QC tests, a standardized format (used at the acceptance test) should be applied, as the results are dependent on the file format used. (orig.)

  2. Intracardiac flow patterns studied by cine MR flow imaging

    International Nuclear Information System (INIS)

    Underwood, S.R.; Firmin, D.N.; Klipstein, R.H.; Rees, R.S.O.; Longmore, D.B.

    1986-01-01

    Velocity mapping by means of cine-MR imaging allows accurate measurement of velocity and flow within the cardiovascular system. A cine display and color coding simplify interpretation. The author have used the technique in a variety of patients to illustrate its potential. Velocity mapping in coronary artery by pass grafts in six patients provided a measure of graft function. Coronary artery velocities were measured in three subjects. Flow was measured through defects in the atrial septum, the ventricular septum, and a Gerbode defect. Velocity was reduced distal to coarctation of the aorta and was increased at the level of a partial venous occlusion by thrombosis. In a patient with isomerism, velocity mapping in the central vessels aided interpretation. Cine-MR imaging velocity mapping combined with conventional imaging yields important functional information on the cardiovascular system

  3. Cine magnetic resonance imaging in congenital heart disease

    International Nuclear Information System (INIS)

    Akagi, Teiji; Kiyomatsu, Yumi; Kato, Hirohisa; Abe, Toushi; Nishimura, Hiroshi; Ohtake, Hisashi; Eto, Takaharu.

    1989-01-01

    Cine magnetic resonance imaging (MRI) was performed in 33 patients aged 19 days to 18 years (mean 5.1 years), who had congenital heart disease comfirmed at echocardiography or angiography. Prior to cine MRI, gated MRI with spin echo (SE) sequence was perfomed to evaluate cardiac structure. Cine MRI was demonstrated by fast low fip angle shot imaging technique with a 30 deg flip angle, 15 msec echo time, 30∼40 msec pulse repetition time, and 128X128 acquisition matrix. Abnormalities of cardiac structure were extremely well defined in all patients. Intracardiac and intravasucular blood flow were visualized with high signal intensity area, whereas ventricular filling flow and left to right shunt flow through ventricular septal defect and atrial septal defect were visualized with low signal intensity area. However, in the patients who had severe congestive heart failure or respiratory arrhythmia, the good recording of cine MRI was not obtained because of artifacts. Gated MRI with SE sequence provides excellent visualization of fine structures, and cine MRI can provide high spatial resolution imaging of flow dynamic in a variety of congenital heart disease, noninvasively. (author)

  4. Dobutamine cine magnetic resonance imaging after myocardial infarction

    International Nuclear Information System (INIS)

    Giovagnoni, A.; Ligabue, G.; Romagnoli, R.; Reggio Emilia Univ., Reggio Emilia; Rossi, R.; Muia, N.; Modena, M.G.; Reggio Emilia Univ.

    1999-01-01

    Dobutamine Cine MRI is a new diagnostic imaging technique in the pretreatment (revascularization) assessment of myocardial infarction patients. In this issue are reported the result of a comparative study of the diagnostic yield of dobutamine Cine MRI with that of stress echocardiography in the assessment of viable myocardium. A new method for analysis of Cine MR images, employing digital subtraction, aimed at decreasing subjectivity in the quantitative assessment of myocardial wall thickening. Twenty-six patients (21 men and 5 women) with a history of myocardial infarction who were scheduled for revascularization were submitted to stress echocardiography and dobutamine Cine MRI to evaluate contractile recovery of the segments considered akinetic or hypo kinetic at baseline echocardiography. Dobutamine was administered in growing doses (5, 10, 15γ/kg/min). 16 segments of the left ventricle in each patient were considered. In the 416 segments studied, it was found that 307 normo kinetic, 64 scarred and 45 viable segments with stress echocardiography, versus 302 normo kinetic, 83 scarred and 31 viable segments with dobutamine MRI. Three months after revascularization 15 patients were examined to check contractile recovery of the segments considered as viable. Echocardiography had 79% sensitivity and 97% specificity, while Cine MRI had 96% and 86%, respectively. In patients with anteroseptal wall myocardial infarction stress echocardiography had 75% sensitivity and 97% specificity. Echocardiography permits to distinguish viable myocardium and scarred myocardial tissue with good sensitivity and specificity, but Cine MRI performs better. Cine MRI has much higher sensitivity than stress echocardiography and thus makes the technique of choice to evaluate viable myocardium in these sites. The digital subtraction technique is as accurate as manual measurements, but reduces the error rate and permits quicker evaluation, particularly in subendocardial thickening [it

  5. Use of local noise power spectrum and wavelet analysis in quantitative image quality assurance for EPIDs

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Soyoung [Department of Radiation Oncology, University Hospitals Case and Medical Center, Cleveland, Ohio 44106 (United States); Yan, Guanghua; Bassett, Philip; Samant, Sanjiv, E-mail: samant@ufl.edu [Department of Radiation Oncology, University of Florida College of Medicine, Gainesville, Florida 32608 (United States); Gopal, Arun [Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, Maryland 21201 (United States)

    2016-09-15

    Purpose: To investigate the use of local noise power spectrum (NPS) to characterize image noise and wavelet analysis to isolate defective pixels and inter-subpanel flat-fielding artifacts for quantitative quality assurance (QA) of electronic portal imaging devices (EPIDs). Methods: A total of 93 image sets including custom-made bar-pattern images and open exposure images were collected from four iViewGT a-Si EPID systems over three years. Global quantitative metrics such as modulation transform function (MTF), NPS, and detective quantum efficiency (DQE) were computed for each image set. Local NPS was also calculated for individual subpanels by sampling region of interests within each subpanel of the EPID. The 1D NPS, obtained by radially averaging the 2D NPS, was fitted to a power-law function. The r-square value of the linear regression analysis was used as a singular metric to characterize the noise properties of individual subpanels of the EPID. The sensitivity of the local NPS was first compared with the global quantitative metrics using historical image sets. It was then compared with two commonly used commercial QA systems with images collected after applying two different EPID calibration methods (single-level gain and multilevel gain). To detect isolated defective pixels and inter-subpanel flat-fielding artifacts, Haar wavelet transform was applied on the images. Results: Global quantitative metrics including MTF, NPS, and DQE showed little change over the period of data collection. On the contrary, a strong correlation between the local NPS (r-square values) and the variation of the EPID noise condition was observed. The local NPS analysis indicated image quality improvement with the r-square values increased from 0.80 ± 0.03 (before calibration) to 0.85 ± 0.03 (after single-level gain calibration) and to 0.96 ± 0.03 (after multilevel gain calibration), while the commercial QA systems failed to distinguish the image quality improvement between the two

  6. Magnetic resonance (MR) cine imaging of the human heart

    International Nuclear Information System (INIS)

    Waterton, J.C.

    1985-01-01

    A novel approach has been developed for MR cine imaging of the human heart by a modified ECG-gated 2DFT method. A pulse sequence has been devised to minimise the effects of saturation which can be anticipated in sequences that require rapid pulsing. Five frames are produced at the same anatomical level at predetermined intervals during the cardiac cycle. The total time taken to achieve this data is 8 minutes. Additional frames can be interleaved by repeating the sequence with an ECG-gated delay. The anatomical sections, which can be in any orthogonal plane, are then displayed as a cine loop. Cine display in the coronal plane has been used to examine 10 volunteers and 12 patients. In addition to the morphological feature displayed in single slice ECG-gated imaging, areas of dyskinesia can be detected and subjective estimates have been made of left ventricular function. (author)

  7. Evaluation of usefulness of portal image using Electronic Portal Imaging Device (EPID) in the patients who received pelvic radiation therapy

    International Nuclear Information System (INIS)

    Kim, Woo Chul; Kim, Heon Jong; Park, Seong Young; Cho, Young Kap; Loh, John J. K.; Park, Won; Suh, Chang Ok; Kim, Gwi Eon

    1998-01-01

    To evaluate the usefulness of electronic portal imaging device through objective compare of the images acquired using an EPID and a conventional port film. From Apr. to Oct. 1997, a total of 150 sets of images from 20 patients who received radiation therapy in the pelvis area were evaluated in the Inha University Hospital and Severance Hospital. A dual image recording technique was devised to obtain both electronic portal images and port film images simultaneously with one treatment course. We did not perform double exposure. Five to ten images were acquired from each patient. All images were acquired from posteroanterior (PA) view except images from two patients. A dose rate of 100-300 MU/min and a 10-MV X-ray beam were used and 2-10 MUs were required to produce a verification image during treatment. Kodak diagnostic film with metal/film imaging cassette which was located on the top of the EPID detector was used for the port film. The source to detector distance was 140 cm. Eight anatomical landmarks (pelvic brim, sacrum, acetabulum, iliopectineal line, symphysis, ischium, obturator foramen, sacroiliac joint) were assessed. Four radiation oncologist joined to evaluate each image. The individual landmarks in the port film or in the EPID were rated-very clear (1), clear (2), visible (3), notclear (4), not visible (5). Using an video camera based EPID system, there was no difference of image quality between no enhanced EPID images and port film images. However, when we provided some change with window level for the portal image, the visibility of the sacrum and obturator foramen was improved in the portal images than in the port film images. All anatomical landmarks were more visible in the portal images than in the port film when we applied the CLAHE mode enhancement. The images acquired using an matrix ion chamber type EPID were also improved image quality after window level adjustment. The quality of image acquired using an electronic portal imaging device was

  8. Fundamental Tongue Motions for Trumpet Playing: A Study Using Cine Magnetic Resonance Imaging (Cine MRI).

    Science.gov (United States)

    Furuhashi, Hiroko; Chikui, Toru; Inadomi, Daisuke; Shiraishi, Tomoko; Yoshiura, Kazunori

    2017-12-01

    Though the motions of structures outside the mouth in trumpet performance have been reported, the dynamics of intraoral structures remain unelucidated. This study explored the tongue's movement in trumpet playing using cine magnetic resonance imaging (cine MRI) and demonstrated the effects of intraoral anatomical structures on changes in pitch and dynamics. Cine MRI was applied to 18 trumpet players, who were divided into two groups (7 beginner, 11 advanced) based on their ability to play a certain high note. They were instructed to play a custom-made MRI-compatible simulated trumpet. Pitch-change tasks and dynamics-change tasks were assigned. The positions of the anatomical points and intraoral areas were identified on outlined images, and the changes associated with each task were evaluated. A forward and upward projection of the tongue was observed in the production of higher pitches, and there were no significant differences in all areas. In louder dynamics, a backward and downward bending of the tongue occurred, the tongue area became smaller (pcine MRI that certain tongue movements were associated with each task. Tongue protrusion in the production of higher pitch and bending in louder dynamics can be rationalized using acoustics theory and the movements of anatomical structures. These findings seem to be consistent regardless of the player's proficiency.

  9. Ghosting effect in Siemens electronic portal imaging devices (EPIDs) for step and shoot IMRT dosimetry

    International Nuclear Information System (INIS)

    Deshpande, S.; Vial, P.; Goozee, G.; Holloway, L.

    2010-01-01

    Full text: To assess the ghosting effect of a Siemens EPID (Optivue 1000: while acquiring IMRT fluence with step and shoot delivery. Six sets of segmented fields with 1,2,3,5, J( and 20 monitor units (MU) per segment were designed. Each set consisted of ten segments of equal MU and field size (J 0 x 10 cm 2 ) Standard single fields (non-segmented) of the same total MU as the segmented fields were also created (10-200 MU). EPID images for these fields were acquired with multi-frame acquisition mode. The integrated EPID response was determined as the mean central 20 x 21 pixel readout multiplied by the number of frames. The same fields wen measured with an ionization chamber at a depth of dose maximum in, solid water phantom. The total signal measured from the segmented fields was compared to the corresponding non-segmented fields. The ratio of EPID response between segmented and non-segmented delivery indicates an under-response for segmented fields by 5, 4, 2.5 and 2% for 1,2,3, and 5 MU per segment exposures respectively compared to ionisation chamber response (se Fig. I). The ratio was within 2% for 5 MU per segment and above. Th error bar in Fig. I indicate the intra-segment response variation. The Siemens EPID exhibited significant ghosting effect and variation in response for small M U segments. EPID dosimetry ( IMRT fields with less than 5 MU per segment requires corrections t maintain dose calibration accuracy to within 2%. (author)

  10. Image processing in cine-scintigraphy. Organization of the CINE 200 system

    International Nuclear Information System (INIS)

    Tournier, E.; Axelrad, C.; Monge, J.; Beccia, A.; Bouvier, A.

    1975-01-01

    The CINE 200 system, although using a small computer to reduce costs, is shown to be a highly efficient instrument for the processing of scintigraphic images. This was achieved by creating an acquisition and display system branched directly onto the memory, reducing the processing times by the writing of fast micro-programmes, and cutting down the place occupied by the programmes in the live memory by a good organization of these programmes. This internal organization of the system, the insertion and role of the programmes, the data organization and the working procedure are described briefly [fr

  11. Evaluation of mitral regurgitation by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Kobayashi, Shiro; Kishi, Naohiro; Kumai, Toshihiko

    1993-01-01

    Valvular regurgitation can be detected as a region of signal loss ('flow void') by cardiac cine magnetic resonance (MR) imaging. Cine MR images of 36 patients with mitral regurgitation caused by mitral valve prolapse (MVP) and non-MVP were evaluated and compared with color Doppler flow images. The direction, distance, area and timing of flow void were detected in all patients in several different planes at mitral level with a 0.5 Tesla superconducting magnet by an ECG-gated fast field echo technique. In 23 of the 36 patients contiguous multiple transaxial images were also obtained to calculate the volumes of flow void and the left atrium. The direction of flow void tended to deviate to the opposite side within the left atrium in MVP. The frame showing maximal area of flow void was demonstrated in mid-systole in 24 of the 36 patients (67%). The distance, area and volume of flow void were concordant with the grade from color Doppler flow images. The volumes of flow void and the left atrium correlated (n=19, r=0.74, p<0.05) in MVP. In conclusion, cine MR images in several different planes or contiguous multiple slices are useful in determining spatial orientation and the extent and timing of mitral regurgitation noninvasively. Furthermore, calculation of the volume of flow void enables the assessment of the semiquantitation of mitral regurgitation. (author)

  12. Evaluation of aortic distensibility using cine MR imaging

    International Nuclear Information System (INIS)

    Takahashi, Masashi; Murata, Kiyoshi; Mori, Masayuki

    1992-01-01

    Regional aortic distensibility was measured noninvasively using cine MR imaging to determine whether it correlates with aging or risk of coronary artery disease (CAD). Twenty-two patients with CAD confirmed by angiography and 33 without CAD underwent cine MR imaging. Thirty-two sequential images were obtained in one cardiac cycle. The cross sectional area of the descending thoracic aorta was measured on both diastolic (A) and systolic (A') images. Aortic distensibility (A.D.) was calculated from the following equation: A.D.=(A'-A)/A/ΔP.(ΔP represents pulse pressure). Results of a simple regression analysis showed that decreased regional aortic distensibility was significantly correlated with the age of the patients (r=-0.762, p<0.005). In addition, the parameter was lower in patients with CAD than those without CAD (p<0.05). In conclusion, regional aortic distensibility derived from cine MR imaging is a useful parameter to evaluate not only aging but also pathological atherosclerosis of the aorta. In addition, this parameter might have some potential to evaluate the prediction of CAD in patients with normal serum cholesterol. (author)

  13. WE-DE-BRA-06: Evaluation of the Imaging Performance of a Novel Water-Equivalent EPID

    Energy Technology Data Exchange (ETDEWEB)

    Blake, SJ [School of Physics, The University of Sydney, Sydney, NSW (Australia); The Ingham Institute, Liverpool, NSW (Australia); Cheng, J; Atakaramians, S; Kuncic, Z [School of Physics, The University of Sydney, Sydney, NSW (Australia); Vial, P [School of Physics, The University of Sydney, Sydney, NSW (Australia); The Ingham Institute, Liverpool, NSW (Australia); Department of Medical Physics, Liverpool & Macarthur Cancer Therapy Centres, Liverpool, NSW (Australia); Lu, M [Perkin-Elmer Medical Imaging, Santa Clara, California (United States); Meikle, S [Faculty of Health Sciences and Brain and Mind Centre, The University of Sydney, Sydney, NSW (Australia)

    2016-06-15

    Purpose: To evaluate the megavoltage imaging performance of a novel, water-equivalent electronic portal imaging device (EPID) developed for simultaneous imaging and dosimetry applications in radiotherapy. Methods: A novel EPID prototype based on active matrix flat panel imager technology has been developed by our group and previously reported to exhibit a water-equivalent dose response. It was constructed by replacing all components above the photodiode detector in a standard clinical EPID (including the copper plate and phosphor screen) with a 15 × 15 cm{sup 2} array of plastic scintillator fibers. Individual fibers measured 0.5 × 0.5 × 30 mm{sup 3}. Spatial resolution was evaluated experimentally relative to that of a standard EPID with the thin slit technique to measure the modulation transfer function (MTF) for 6 MV x-ray beams. Monte Carlo (MC) EPID models were used to benchmark simulated MTFs against the measurements. The zero spatial frequency detective quantum efficiency (DQE(0)) was simulated for both EPID configurations and a preliminary optimization of the prototype was performed by evaluating DQE(0) as a function of fiber length up to 50 mm. Results: The MC-simulated DQE(0) for the prototype EPID configuration was ∼7 times greater than that of the standard EPID. The prototype’s DQE(0) also increased approximately linearly with fiber length, from ∼1% at 5 mm length to ∼11% at 50 mm length. The standard EPID MTF was greater than the prototype EPID’s for all spatial frequencies, reflecting the trade off between x-ray detection efficiency and spatial resolution with thick scintillators. Conclusion: This study offers promising evidence that a water-equivalent EPID previously demonstrated for radiotherapy dosimetry may also be used for radiotherapy imaging applications. Future studies on optimising the detector design will be performed to develop a next-generation prototype that offers improved megavoltage imaging performance, with the aim to at

  14. WE-DE-BRA-06: Evaluation of the Imaging Performance of a Novel Water-Equivalent EPID

    International Nuclear Information System (INIS)

    Blake, SJ; Cheng, J; Atakaramians, S; Kuncic, Z; Vial, P; Lu, M; Meikle, S

    2016-01-01

    Purpose: To evaluate the megavoltage imaging performance of a novel, water-equivalent electronic portal imaging device (EPID) developed for simultaneous imaging and dosimetry applications in radiotherapy. Methods: A novel EPID prototype based on active matrix flat panel imager technology has been developed by our group and previously reported to exhibit a water-equivalent dose response. It was constructed by replacing all components above the photodiode detector in a standard clinical EPID (including the copper plate and phosphor screen) with a 15 × 15 cm 2 array of plastic scintillator fibers. Individual fibers measured 0.5 × 0.5 × 30 mm 3 . Spatial resolution was evaluated experimentally relative to that of a standard EPID with the thin slit technique to measure the modulation transfer function (MTF) for 6 MV x-ray beams. Monte Carlo (MC) EPID models were used to benchmark simulated MTFs against the measurements. The zero spatial frequency detective quantum efficiency (DQE(0)) was simulated for both EPID configurations and a preliminary optimization of the prototype was performed by evaluating DQE(0) as a function of fiber length up to 50 mm. Results: The MC-simulated DQE(0) for the prototype EPID configuration was ∼7 times greater than that of the standard EPID. The prototype’s DQE(0) also increased approximately linearly with fiber length, from ∼1% at 5 mm length to ∼11% at 50 mm length. The standard EPID MTF was greater than the prototype EPID’s for all spatial frequencies, reflecting the trade off between x-ray detection efficiency and spatial resolution with thick scintillators. Conclusion: This study offers promising evidence that a water-equivalent EPID previously demonstrated for radiotherapy dosimetry may also be used for radiotherapy imaging applications. Future studies on optimising the detector design will be performed to develop a next-generation prototype that offers improved megavoltage imaging performance, with the aim to at least

  15. Using an EPID for patient-specific VMAT quality assurance

    International Nuclear Information System (INIS)

    Bakhtiari, M.; Kumaraswamy, L.; Bailey, D. W.; Boer, S. de; Malhotra, H. K.; Podgorsak, M. B.

    2011-01-01

    Purpose: A patient-specific quality assurance (QA) method was developed to verify gantry-specific individual multileaf collimator (MLC) apertures (control points) in volumetric modulated arc therapy (VMAT) plans using an electronic portal imaging device (EPID). Methods: VMAT treatment plans were generated in an Eclipse treatment planning system (TPS). DICOM images from a Varian EPID (aS1000) acquired in continuous acquisition mode were used for pretreatment QA. Each cine image file contains the grayscale image of the MLC aperture related to its specific control point and the corresponding gantry angle information. The TPS MLC file of this RapidArc plan contains the leaf positions for all 177 control points (gantry angles). In-house software was developed that interpolates the measured images based on the gantry angle and overlays them with the MLC pattern for all control points. The 38% isointensity line was used to define the edge of the MLC leaves on the portal images. The software generates graphs and tables that provide analysis for the number of mismatched leaf positions for a chosen distance to agreement at each control point and the frequency in which each particular leaf mismatches for the entire arc. Results: Seven patients plans were analyzed using this method. The leaves with the highest mismatched rate were found to be treatment plan dependent. Conclusions: This in-house software can be used to automatically verify the MLC leaf positions for all control points of VMAT plans using cine images acquired by an EPID.

  16. Two-dimensional sonographic cine imaging improves confidence in the initial evaluation of the fetal heart.

    Science.gov (United States)

    Poole, Patricia Sims; Chung, Romy; Lacoursiere, Yvette; Palmieri, Carolina Rossi; Hull, Andrew; Engelkemier, Dawn; Rochelle, Michele; Trivedi, Neha; Pretorius, Dolores H

    2013-06-01

    Initial screening sonography of the fetal heart with static images is often inadequate, resulting in repeated imaging or failure to detect abnormalities. We hypothesized that the addition of short cine clips would reduce the need for repeated imaging. Two-dimensional (2D) static sonograms and short 2D cine clips of the 4-chamber view and left and right ventricular outflow tracts were obtained from 342 patients with gestational ages of greater than 16 weeks. A diagnostic radiologist and a perinatologist retrospectively reviewed the static and cine images independently and graded them as normal, abnormal, or suboptimal. A statistically significant increase in the number of structures called normal was seen when 2D cine clips were added to static imaging for both observers (P cine images versus 61.9% with static images alone, whereas the perinatologist recorded 68.1% as normal versus 58.8%, respectively. The radiologist called 77.8% of structures normal with cine images only versus 61.9% with static images only (P cine images alone (38.9%) versus static images alone (58.8%). The use of cine loops alone resulted in no significant increase in the ability to clear the heart as normal. The maternal body mass index was inversely associated with the ability to clear structures when 2D cine images were added to static images (P cine clips to standard 2D static imaging of the fetal heart significantly improves the number of structures cleared as normal. Two-dimensional cine clips are easily obtained, add little time to a study, and require minimal archival space.

  17. Comparisons of images simultaneously documented by digital subtraction coronary arteriography and cine coronary arteriography

    International Nuclear Information System (INIS)

    Kimura, Koji; Takamiya, Makoto; Yamamoto, Kazuo; Ohta, Mitsushige; Naito, Hiroaki

    1988-01-01

    Using an angiography apparatus capable of simultaneously processing digital subtraction angiograms and cine angiograms, the diagnostic capabilities of both methods for the coronary arteries (DSCAG and Cine-CAG) were compared. Twenty stenotic lesions of the coronary arteries of 11 patients were evaluated using both modalities. The severity of stenosis using DSCAG with a 512x512x8 bit matrix was semiautomatically measured on the cathode ray tube (CRT) based on enlarged images on the screen of a Vanguard cine projector which were of the same size as those of or 10 times larger than images of Cine-CAG. The negative and positive hard copies of DSCAG images were also compared with those of Cine-CAG. The correlation coefficients of the severity of stenosis by DSCAG and Cine-CAG were as follows: (1) the same size DSCAG images on CRT to Cine-CAG, 0.95, (2) 10 times enlarged DSCAG images on CRT to Cine-CAG, 0.96, and (3) the same size DSCAG images on negative and positive hard copies to Cine-CAG, 0.97. The semiautomatically measured values of 10 times enlarged DSCAG images on CRT and the manually measured values of the same size negative and positive DSCAG images in hard copy closely correlated with the values measured using Cine-CAG. When the liver was superimposed in the long-axis projection, the diagnostic capabilities of DSCAG and Cine-CAG were compared. The materials included 10 left coronary arteriograms and 11 right coronary arteriograms. Diagnostically, DSCAG was more useful than Cine-CAG in the long-axis projection. (author)

  18. Physiologic AV valvular insufficiency in cine MR imaging

    Energy Technology Data Exchange (ETDEWEB)

    Jang, Yoon Hyung; Kang, Eun Joo; Baik, Seung Kug; Ahn, Woo Hyun; Choi, Han Yong; Kim, Bong Gi [Wallace Memorial Baptist Hospital, Pusan (Korea, Republic of)

    1994-05-15

    To give a help in the interpretation of cardiac cine-MR examination, the extent, shape, and timing of appearance of signal void regions near atrioventricular(A-V) valve prospectively evaluate in the healthy population. Using an axial gradient-echo technique with small flip angle, repetition time(TR) of 36 msec and echo time(TE) of 22 msec, 20 volunteers without known valvular abnormalities undertook cardiac cine-MR imaging including atrioventricular valve areas. Transient signal void was observed within the near the tricuspid(13/20 = 65%) and mitral valves(9/20 = 45%), respectively, which is so called {sup p}hysioloic atrioventricular valvular insufficiency{sup .} Eight subjects revealed the signal void areas near both tricuspid and mitral valves but, 5 subjects did not show any evidence of physiologic insufficiency. This physiologic condition does not extend more than 1 cm proximal to A-V valve plane and is generally observed only during early systole. Its morphology is semilunar or triangular configuration with the base to the valve plane in most cases of normal tricuspid insufficiency and small globular appearance in most cases of normal mitral insufficiency. Awareness of normal signal void areas near the A-V valve and their characteristics is critical in the interpretation of cardiac cine MR examinations and maybe helpful in the study of the normal cardiac physiology.

  19. Physiologic AV valvular insufficiency in cine MR imaging

    International Nuclear Information System (INIS)

    Jang, Yoon Hyung; Kang, Eun Joo; Baik, Seung Kug; Ahn, Woo Hyun; Choi, Han Yong; Kim, Bong Gi

    1994-01-01

    To give a help in the interpretation of cardiac cine-MR examination, the extent, shape, and timing of appearance of signal void regions near atrioventricular(A-V) valve prospectively evaluate in the healthy population. Using an axial gradient-echo technique with small flip angle, repetition time(TR) of 36 msec and echo time(TE) of 22 msec, 20 volunteers without known valvular abnormalities undertook cardiac cine-MR imaging including atrioventricular valve areas. Transient signal void was observed within the near the tricuspid(13/20 = 65%) and mitral valves(9/20 = 45%), respectively, which is so called p hysioloic atrioventricular valvular insufficiency . Eight subjects revealed the signal void areas near both tricuspid and mitral valves but, 5 subjects did not show any evidence of physiologic insufficiency. This physiologic condition does not extend more than 1 cm proximal to A-V valve plane and is generally observed only during early systole. Its morphology is semilunar or triangular configuration with the base to the valve plane in most cases of normal tricuspid insufficiency and small globular appearance in most cases of normal mitral insufficiency. Awareness of normal signal void areas near the A-V valve and their characteristics is critical in the interpretation of cardiac cine MR examinations and maybe helpful in the study of the normal cardiac physiology

  20. New developments in imaging: Sonography, cine-CT, MRI

    International Nuclear Information System (INIS)

    Otto, R.J.; Higgins, C.B.

    1987-01-01

    The book can be conveniently subdivided into three sections: the first on magnetic resonance imaging the second on cine-computed tomography and the third on advances in ultrasound (US). The MR imaging section includes two chapters: the first on indications for MR in abdominal disease (a cookbook layout of indications for MR imaging versus CT) and the second on MR imaging of the heart. There are also chapters on MR imaging and US in the pelvis, contrast agent principles, and a chapter on imaging renal tumors. The third section, on US, contains chapters on the liver and gastrointenstinal disease, interventional US sonography during neurosurgery, state-of-the-art echocardiography. Doppler flow imaging, contrast media for sonography, endometrial sonography, and high-resolution US in the first trimester. The final chapter is presented as a scientific paper rather than as a chapter in a book and has no illustrations

  1. A Technique for Generating Volumetric Cine-Magnetic Resonance Imaging

    International Nuclear Information System (INIS)

    Harris, Wendy; Ren, Lei; Cai, Jing; Zhang, You; Chang, Zheng; Yin, Fang-Fang

    2016-01-01

    Purpose: The purpose of this study was to develop a techique to generate on-board volumetric cine-magnetic resonance imaging (VC-MRI) using patient prior images, motion modeling, and on-board 2-dimensional cine MRI. Methods and Materials: One phase of a 4-dimensional MRI acquired during patient simulation is used as patient prior images. Three major respiratory deformation patterns of the patient are extracted from 4-dimensional MRI based on principal-component analysis. The on-board VC-MRI at any instant is considered as a deformation of the prior MRI. The deformation field is represented as a linear combination of the 3 major deformation patterns. The coefficients of the deformation patterns are solved by the data fidelity constraint using the acquired on-board single 2-dimensional cine MRI. The method was evaluated using both digital extended-cardiac torso (XCAT) simulation of lung cancer patients and MRI data from 4 real liver cancer patients. The accuracy of the estimated VC-MRI was quantitatively evaluated using volume-percent-difference (VPD), center-of-mass-shift (COMS), and target tracking errors. Effects of acquisition orientation, region-of-interest (ROI) selection, patient breathing pattern change, and noise on the estimation accuracy were also evaluated. Results: Image subtraction of ground-truth with estimated on-board VC-MRI shows fewer differences than image subtraction of ground-truth with prior image. Agreement between normalized profiles in the estimated and ground-truth VC-MRI was achieved with less than 6% error for both XCAT and patient data. Among all XCAT scenarios, the VPD between ground-truth and estimated lesion volumes was, on average, 8.43 ± 1.52% and the COMS was, on average, 0.93 ± 0.58 mm across all time steps for estimation based on the ROI region in the sagittal cine images. Matching to ROI in the sagittal view achieved better accuracy when there was substantial breathing pattern change. The technique was robust against

  2. A Technique for Generating Volumetric Cine-Magnetic Resonance Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Harris, Wendy [Medical Physics Graduate Program, Duke University, Durham, North Carolina (United States); Ren, Lei, E-mail: lei.ren@duke.edu [Medical Physics Graduate Program, Duke University, Durham, North Carolina (United States); Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina (United States); Cai, Jing [Medical Physics Graduate Program, Duke University, Durham, North Carolina (United States); Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina (United States); Zhang, You [Medical Physics Graduate Program, Duke University, Durham, North Carolina (United States); Chang, Zheng; Yin, Fang-Fang [Medical Physics Graduate Program, Duke University, Durham, North Carolina (United States); Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina (United States)

    2016-06-01

    Purpose: The purpose of this study was to develop a techique to generate on-board volumetric cine-magnetic resonance imaging (VC-MRI) using patient prior images, motion modeling, and on-board 2-dimensional cine MRI. Methods and Materials: One phase of a 4-dimensional MRI acquired during patient simulation is used as patient prior images. Three major respiratory deformation patterns of the patient are extracted from 4-dimensional MRI based on principal-component analysis. The on-board VC-MRI at any instant is considered as a deformation of the prior MRI. The deformation field is represented as a linear combination of the 3 major deformation patterns. The coefficients of the deformation patterns are solved by the data fidelity constraint using the acquired on-board single 2-dimensional cine MRI. The method was evaluated using both digital extended-cardiac torso (XCAT) simulation of lung cancer patients and MRI data from 4 real liver cancer patients. The accuracy of the estimated VC-MRI was quantitatively evaluated using volume-percent-difference (VPD), center-of-mass-shift (COMS), and target tracking errors. Effects of acquisition orientation, region-of-interest (ROI) selection, patient breathing pattern change, and noise on the estimation accuracy were also evaluated. Results: Image subtraction of ground-truth with estimated on-board VC-MRI shows fewer differences than image subtraction of ground-truth with prior image. Agreement between normalized profiles in the estimated and ground-truth VC-MRI was achieved with less than 6% error for both XCAT and patient data. Among all XCAT scenarios, the VPD between ground-truth and estimated lesion volumes was, on average, 8.43 ± 1.52% and the COMS was, on average, 0.93 ± 0.58 mm across all time steps for estimation based on the ROI region in the sagittal cine images. Matching to ROI in the sagittal view achieved better accuracy when there was substantial breathing pattern change. The technique was robust against

  3. Single-breath-hold 3-D CINE imaging of the left ventricle using Cartesian sampling.

    Science.gov (United States)

    Wetzl, Jens; Schmidt, Michaela; Pontana, François; Longère, Benjamin; Lugauer, Felix; Maier, Andreas; Hornegger, Joachim; Forman, Christoph

    2018-02-01

    Our objectives were to evaluate a single-breath-hold approach for Cartesian 3-D CINE imaging of the left ventricle with a nearly isotropic resolution of [Formula: see text] and a breath-hold duration of [Formula: see text]19 s against a standard stack of 2-D CINE slices acquired in multiple breath-holds. Validation is performed with data sets from ten healthy volunteers. A Cartesian sampling pattern based on the spiral phyllotaxis and a compressed sensing reconstruction method are proposed to allow 3-D CINE imaging with high acceleration factors. The fully integrated reconstruction uses multiple graphics processing units to speed up the reconstruction. The 2-D CINE and 3-D CINE are compared based on ventricular function parameters, contrast-to-noise ratio and edge sharpness measurements. Visual comparisons of corresponding short-axis slices of 2-D and 3-D CINE show an excellent match, while 3-D CINE also allows reformatting to other orientations. Ventricular function parameters do not significantly differ from values based on 2-D CINE imaging. Reconstruction times are below 4 min. We demonstrate single-breath-hold 3-D CINE imaging in volunteers and three example patient cases, which features fast reconstruction and allows reformatting to arbitrary orientations.

  4. Prostate gland motion assessed with cine-magnetic resonance imaging (cine-MRI)

    International Nuclear Information System (INIS)

    Ghilezan, Michel J.; Jaffray, David A.; Siewerdsen, Jeffrey H.; Herk, Marcel van; Shetty, Anil; Sharpe, Michael B.; Zafar Jafri, Syed; Vicini, Frank A.; Matter, Richard C.; Brabbins, Donald S.; Martinez, Alvaro A.

    2005-01-01

    Purpose: To quantify prostate motion during a radiation therapy treatment using cine-magnetic resonance imaging (cine-MRI) for time frames comparable to that expected in an image-guided radiation therapy treatment session (20-30 min). Materials and Methods: Six patients undergoing radiation therapy for prostate cancer were imaged on 3 days, over the course of therapy (Weeks 1, 3, and 5). Four hundred images were acquired during the 1-h MRI session in 3 sagittal planes through the prostate at 6-s intervals. Eleven anatomic points of interest (POIs) have been used to characterize prostate/bony pelvis/abdominal wall displacement. Motion traces and standard deviation for each of the 11 POIs have been determined. The probability of displacement over time has also been calculated. Results: Patients were divided into 2 groups according to rectal filling status: full vs. empty rectum. The displacement of POIs (standard deviation) ranged from 0.98 to 1.72 mm for the full-rectum group and from 0.68 to 1.04 mm for the empty-rectum group. The low standard deviations in position (2 mm or less) would suggest that these excursions have a low frequency of occurrence. The most sensitive prostate POI to rectal wall motion was the midposterior with a standard deviation of 1.72 mm in the full-rectum group vs. 0.79 mm in the empty-rectum group (p 0.0001). This POI has a 10% probability of moving more than 3 mm in a time frame of ∼1 min if the rectum is full vs. ∼20 min if the rectum is empty. Conclusion: Motion of the prostate and seminal vesicles during a time frame similar to a standard treatment session is reduced compared to that reported in interfraction studies. The most significant predictor for intrafraction prostate motion is the status of rectal filling. A prostate displacement of <3 mm (90%) can be expected for the 20 min after the moment of initial imaging for patients with an empty rectum. This is not the case for patients presenting with full rectum. The determination

  5. Usefulness of true FISP cine MR imaging in patients with poor cardiac function

    Energy Technology Data Exchange (ETDEWEB)

    Sakuma, Toshiharu; Yamada, Naoaki; Motooka, Makoto; Enomoto, Naoyuki; Maeshima, Isamu; Matsuda, Kazuhide; Urayama, Shinichi; Ikeo, Miki [National Cardiovascular Center, Suita, Osaka (Japan)

    2002-01-01

    This study was done to assess the value of True FISP cine in patients with poor cardiac function. True FISP cine and FLASH cine imaging were performed on a 1.5 T machine. Both short axis and horizontal long axis imaging sections were used. The imaging sections used a Matrix (120 x 128), FOV (24 x 32 cm), and had a slice thickness of 8 mm. The imaging time for True FISP cine was 8 heart beats and 17 heart beats for FLASH cine. The contrast-to-noise ratio between the blood and myocardium (CNR) was measured at enddiastole and endsystole. The subjects in the study were 10 healty volunteers (average age 26.5{+-}3.2 years) and 12 patients with hypofunction (average age 53.9{+-}13.2 years). In the volunteers, the CNR of the short axis imaging was similar in both True FISP (24.6{+-}3.7) and FLASH (23.4{+-}5.9). In the patients with poor cardiac function however, the CNR of True FISP was larger than FLASH in both the short and long axis. In the short axis (22.7{+-}6.1 vs. 17.9{+-}5.3, P<0.01) and in the long axis (17.4{+-}4.3 vs. 9.3{+-}4.0, P<0.01). We conclude that True FISP cine has a higher contrast in a shorter imaging time than FLASH cine. True FISP cine is especially useful in patients with poor cardiac function. (author)

  6. A study of intraluminal flow in patients with aortic disease by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Ohnishi, Shusaku; Fukui, Sugao; Kusuoka, Hideo; Kitabatake, Akira; Kamada, Takenobu.

    1990-01-01

    To evaluate the usefulness of cine magnetic resonance (MR) imaging in the diagnosis of aortic disease, we applied cine MR sequences with conventional spin echo (SE) sequence to 15 patients with aortic aneurysm and 5 patients with aortic dissection. ECG gated multi slice SE images (single echo) were obtained in transverse plane and the planes along the long axis of aorta. TE of SE sequence ranged from 15 to 32 msec. Cine MR images were obtained in 17 frames, where TR ranged from 30 to 60 msec and TE from 10 to 22 msec. We got these results as follows. For aortic aneurysm, SE images showed aneurysmal dilatation in thoracic aorta in 9 cases, and in abdominal aorta in 6 cases. Cine MR images showed us the intraluminal flow with high signal intensity. We could differentiate the slow flow image from mural thrombus using SE and cine MR sequences in the same plane. We could evaluate the dynamic flow image within a cavity of aneurysm and in in-flow and out-flow portions of aorta by cine MR imaging. For aortic dissection, SE images delineated the intimal tear and two separated lumens in all cases. We could obtain flow images within true and false lumen by cine MR imaging. The signal intensity within true lumen was higher than that within false lumen. It seemed to be due to the difference of flow velocity and the turbulence within false lumen. Entry to false lumen was seen as the interruption of dissected intima on SE images. Cine MR images clearly depicted the blood flow communication between two lumens through the entry. In 3 of 5 cases the blood flow through the entry was seen in both directions, from true to false lumen in systole and inversely in diastole. Thus, cine MR imaging will add the useful hemodynamic information in the diagnosis of aortic disease by conventional MR imaging. (author)

  7. Spirometer-controlled cine magnetic resonance imaging used to diagnose tracheobronchomalacia in paediatric patients.

    Science.gov (United States)

    Ciet, Pierluigi; Wielopolski, Piotr; Manniesing, Rashindra; Lever, Sandra; de Bruijne, Marleen; Morana, Giovanni; Muzzio, Pier Carlo; Lequin, Maarten H; Tiddens, Harm A W M

    2014-01-01

    Tracheobronchomalacia (TBM) is defined as an excessive collapse of the intrathoracic trachea. Bronchoscopy is the gold standard for diagnosing TBM; however it has major disadvantages, such as general anaesthesia. Cine computed tomography (CT) is a noninvasive alternative used to diagnose TBM, but its use in children is restricted by ionising radiation. Our aim was to evaluate the feasibility of spirometer-controlled cine magnetic resonance imaging (MRI) as an alternative to cine-CT in a retrospective study. 12 children with a mean age (range) of 12 years (7-17 years), suspected of having TBM, underwent cine-MRI. Static scans were acquired at end-inspiration and expiration covering the thorax using a three-dimensional spoiled gradient echo sequence. Three-dimensional dynamic scans were performed covering only the central airways. TBM was defined as a decrease of the trachea or bronchi diameter >50% at end-expiration in the static and dynamic scans. The success rate of the cine-MRI protocol was 92%. Cine-MRI was compared with bronchoscopy or chest CT in seven subjects. TBM was diagnosed by cine-MRI in seven (58%) out of 12 children and was confirmed by bronchoscopy or CT. In four patients, cine-MRI demonstrated tracheal narrowing that was not present in the static scans. Spirometer controlled cine-MRI is a promising technique to assess TBM in children and has the potential to replace bronchoscopy.

  8. Cine-MR imaging of valvular heart disease

    International Nuclear Information System (INIS)

    Park, Jae Hyung; Han, Man Chung; Kim, Chu Wan; Kim, Dae Jin; Kim, Woo Sung; Park, Hyun Wuk; Cho, Zang Hee

    1989-01-01

    Cine-MR imaging was done in 2 normal persons and 9 patients of valvular heart disease with 2.0 T superconducting MR system (Spectro-2000; GoldStar, Korea). The MRI was undertaken using gradient echo technique with small flip angle. Repetition time (TR) was 34 msec, and echo time (TE), 18 msec. In most cases, 20 to 30 frames could be obtained during one cardiac cycle. With normal heart, high signal intensity was identified in the blood filled cardiac chambers. Transient signal void was observed near tricuspid, mitral and aortic valves due to the turbulences induced by valve opening or closure. In 6 cases of mitral valvular disease, mitral valve was identified in all cases. The valvular motion was limited revealing doming toward cardiac apex during diastole. Signal void was evident in pansystolic phase of left ventricle in all cases. Evidence of combined aortic or tricuspid valve disease were also demonstrated, revealing signal void in the corresponding cardiac chambers. Cine-MRI seemed to be not only a good non-invasive diagnostic modality for the valvular heart disease, but also an accurate modality for cardiac functional evaluation

  9. Cine-MR imaging of valvular heart disease

    Energy Technology Data Exchange (ETDEWEB)

    Park, Jae Hyung; Han, Man Chung; Kim, Chu Wan [Seoul National University College of Medicine, Seoul (Korea, Republic of); Kim, Dae Jin; Kim, Woo Sung; Park, Hyun Wuk; Cho, Zang Hee [Korea Advanced Institute of Science and Technology, Daejeon (Korea, Republic of)

    1989-04-15

    Cine-MR imaging was done in 2 normal persons and 9 patients of valvular heart disease with 2.0 T superconducting MR system (Spectro-2000; GoldStar, Korea). The MRI was undertaken using gradient echo technique with small flip angle. Repetition time (TR) was 34 msec, and echo time (TE), 18 msec. In most cases, 20 to 30 frames could be obtained during one cardiac cycle. With normal heart, high signal intensity was identified in the blood filled cardiac chambers. Transient signal void was observed near tricuspid, mitral and aortic valves due to the turbulences induced by valve opening or closure. In 6 cases of mitral valvular disease, mitral valve was identified in all cases. The valvular motion was limited revealing doming toward cardiac apex during diastole. Signal void was evident in pansystolic phase of left ventricle in all cases. Evidence of combined aortic or tricuspid valve disease were also demonstrated, revealing signal void in the corresponding cardiac chambers. Cine-MRI seemed to be not only a good non-invasive diagnostic modality for the valvular heart disease, but also an accurate modality for cardiac functional evaluation.

  10. Measuring linac photon beam energy through EPID image analysis of physically wedged fields

    Energy Technology Data Exchange (ETDEWEB)

    Dawoud, S. M., E-mail: samir.dawoud@leedsth.nhs.uk; Weston, S. J.; Bond, I.; Ward, G. C.; Rixham, P. A.; Mason, J.; Huckle, A. [Department of Medical Physics and Engineering, St. James Institute of Oncology, St. James University Hospital, Leeds LS9 7TF (United Kingdom); Sykes, J. R. [Institute of Medical Physics, School of Physics, The University of Sydney, New South Wales 2006, Australia and Department of Medical Physics and Engineering, St. James Institute of Oncology, St. James University Hospital, Leeds LS9 7TF (United Kingdom)

    2014-02-15

    Purpose: Electronic portal imaging devices (EPIDs) have proven to be useful tools for measuring several parameters of interest in linac quality assurance (QA). However, a method for measuring linac photon beam energy using EPIDs has not previously been reported. In this report, such a method is devised and tested, based on fitting a second order polynomial to the profiles of physically wedged beams, where the metric of interest is the second order coefficientα. The relationship between α and the beam quality index [percentage depth dose at 10 cm depth (PDD{sub 10})] is examined to produce a suitable calibration curve between these two parameters. Methods: Measurements were taken in a water-tank for beams with a range of energies representative of the local QA tolerances about the nominal value 6 MV. In each case, the beam quality was found in terms of PDD{sub 10} for 100 × 100 mm{sup 2} square fields. EPID images of 200 × 200 mm{sup 2} wedged fields were then taken for each beam and the wedge profile was fitted in MATLAB 2010b (The MathWorks, Inc., Natick, MA). α was then plotted against PDD{sub 10} and fitted with a linear relation to produce the calibration curve. The uncertainty in α was evaluated by taking five repeat EPID images of the wedged field for a beam of 6 MV nominal energy. The consistency of measuring α was found by taking repeat measurements on a single linac over a three month period. The method was also tested at 10 MV by repeating the water-tank crosscalibration for a range of energies centered approximately about a 10 MV nominal value. Finally, the calibration curve from the test linac and that from a separate clinical machine were compared to test consistency of the method across machines in a matched fleet. Results: The relationship betweenα and PDD{sub 10} was found to be strongly linear (R{sup 2} = 0.979) while the uncertainty in α was found to be negligible compared to that associated with measuring PDD{sub 10} in the water-tank (

  11. Dobutamine cine magnetic resonance imaging after myocardial infarction; Cine Risonanza Magnetica con dobutamina dopo infarto del miocardio

    Energy Technology Data Exchange (ETDEWEB)

    Giovagnoni, A.; Ligabue, G.; Romagnoli, R. [Modena Univ., Modena (Italy). Scienze Radiologiche, Dipt. di Medicina Interna; Reggio Emilia Univ., Reggio Emilia (Italy). Cattedra di Cardiologia; Rossi, R.; Muia, N.; Modena, M.G. [Modena Univ., Modena (Italy). Dipt. di Scienze Mediche, Oncologiche e Radiologiche; Reggio Emilia Univ. (Italy). Dipt. di Scienze Mediche, Oncologiche e Radiologiche

    1999-12-01

    Dobutamine Cine MRI is a new diagnostic imaging technique in the pretreatment (revascularization) assessment of myocardial infarction patients. In this issue are reported the result of a comparative study of the diagnostic yield of dobutamine Cine MRI with that of stress echocardiography in the assessment of viable myocardium. A new method for analysis of Cine MR images, employing digital subtraction, aimed at decreasing subjectivity in the quantitative assessment of myocardial wall thickening. Twenty-six patients (21 men and 5 women) with a history of myocardial infarction who were scheduled for revascularization were submitted to stress echocardiography and dobutamine Cine MRI to evaluate contractile recovery of the segments considered akinetic or hypo kinetic at baseline echocardiography. Dobutamine was administered in growing doses (5, 10, 15{gamma}/kg/min). 16 segments of the left ventricle in each patient were considered. In the 416 segments studied, it was found that 307 normo kinetic, 64 scarred and 45 viable segments with stress echocardiography, versus 302 normo kinetic, 83 scarred and 31 viable segments with dobutamine MRI. Three months after revascularization 15 patients were examined to check contractile recovery of the segments considered as viable. Echocardiography had 79% sensitivity and 97% specificity, while Cine MRI had 96% and 86%, respectively. In patients with anteroseptal wall myocardial infarction stress echocardiography had 75% sensitivity and 97% specificity. Echocardiography permits to distinguish viable myocardium and scarred myocardial tissue with good sensitivity and specificity, but Cine MRI performs better. Cine MRI has much higher sensitivity than stress echocardiography and thus makes the technique of choice to evaluate viable myocardium in these sites. The digital subtraction technique is as accurate as manual measurements, but reduces the error rate and permits quicker evaluation, particularly in subendocardial thickening

  12. SU-E-T-781: Using An Electronic Portal Imaging Device (EPID) for Correlating Linac Photon Beam Energies

    Energy Technology Data Exchange (ETDEWEB)

    Yaddanapudi, S; Cai, B; Sun, B; Noel, C; Goddu, S; Mutic, S [Washington University School of Medicine, Saint Louis, MO (United States)

    2015-06-15

    Purpose: Electronic portal imaging devices (EPIDs) have proven to be useful for measuring several parameters of interest in linear accelerator (linac) quality assurance (QA). The purpose of this project was to evaluate the feasibility of using EPIDs for determining linac photon beam energies. Methods: Two non-clinical Varian TrueBeam linacs (Varian Medical Systems, Palo Alto, CA) with 6MV and 10MV photon beams were used to perform the measurements. The linacs were equipped with an amorphous silicon based EPIDs (aSi1000) that were used for the measurements. We compared the use of flatness versus percent depth dose (PDD) for predicting changes in linac photon beam energy. PDD was measured in 1D water tank (Sun Nuclear Corporation, Melbourne FL) and the profiles were measured using 2D ion-chamber array (IC-Profiler, Sun Nuclear) and the EPID. Energy changes were accomplished by varying the bending magnet current (BMC). The evaluated energies conformed with the AAPM TG142 tolerance of ±1% change in PDD. Results: BMC changes correlating with a ±1% change in PDD corresponded with a change in flatness of ∼1% to 2% from baseline values on the EPID. IC Profiler flatness values had the same correlation. We observed a similar trend for the 10MV beam energy changes. Our measurements indicated a strong correlation between changes in linac photon beam energy and changes in flatness. For all machines and energies, beam energy changes produced change in the uniformity (AAPM TG-142), varying from ∼1% to 2.5%. Conclusions: EPID image analysis of beam profiles can be used to determine linac photon beam energy changes. Flatness-based metrics or uniformity as defined by AAPM TG-142 were found to be more sensitive to linac photon beam energy changes than PDD. Research funding provided by Varian Medical Systems. Dr. Sasa Mutic receives compensation for providing patient safety training services from Varian Medical Systems, the sponsor of this study.

  13. The diagnostic value of cine-MR imaging in diseases of great vessels

    International Nuclear Information System (INIS)

    Sasaki, Shigeyuki; Yoshida, Hideaki; Matsui, Yoshiro; Sakuma, Makoto; Yasuda, Keihide; Tanabe, Tatsuzo; Chouji, H.

    1990-01-01

    The diagnostic value of cine magnetic resonance imaging (cine-MRI) was evaluated in 10 patients with diseases of great vessels. The parameters necessary to decide the appropriate treatment, such as presence and extension of intimal flap, DeBakey type classification, identification of the entry, differentiation between true and false lumen, and between thrombosis and slow flow were demonstrated in all patients with dissecting aortic aneurysm. However, abdominal aortic branches could not be demonstrated enough by cine-MRI, therefore conventional AOG was necessary to choose the operative procedure in these cases. In patients with thoracic aortic aneurysm (TAA), cine-MRI was valuable in demonstrating both blood flow and thrombus in the lumen of aneurysm, and AOG was thought to be unnecessary in most cases. Cine-MRI is a promising new technique for the evaluation of diseases of great vessels. (author)

  14. Cine-magnetic resonance imaging evaluation of communication between middle cranial fossa arachnoid cysts and cisterns

    International Nuclear Information System (INIS)

    Eguchi, Takahiko; Nikaido, Yuji; Shiomi, Kazuaki; Fujimoto, Takatoshi; Otsuka, Hiroyuki; Takeuchi, Hiroshi; Taoka, Toshiaki.

    1996-01-01

    Cine-magnetic resonance (MR) imaging examinations were performed in 10 patients with middle cranial fossa arachnoid cysts to evaluate communication between the cysts and the normal cerebrospinal fluid (CSF) space. Eight of 10 patients were evaluated by time of flight cine-MR imaging, and two by phase contrast cine-MR imaging. Two patients underwent membranectomy of the cysts, and were evaluated both pre-and postoperatively. Computed tomography cisternography was used to confirm communication between the cysts and the surrounding cisterns. Pulsatile fluid motion within the cysts was present in all patients. However, marked fluid motion and jet flow between the cysts and the surrounding cisterns were only observed in communicating cysts. In the two patients who underwent membranectomy, postoperative examination found greater fluid motion and jet flow not previously present. Cine-MR imaging demonstration of marked pulsatile fluid motion accompanied by jet flow suggests that a cyst communicates with the normal CSF space. (author)

  15. Cine magnetic resonance imaging for evaluation of cardiac structure and flow dynamics in congenital heart disease

    International Nuclear Information System (INIS)

    Akagi, Teiji; Kiyomatsu, Yumi; Ohara, Nobutoshi; Takagi, Junichi; Sato, Noboru; Kato, Hirohisa; Eto, Takaharu.

    1989-01-01

    Cine magnetic resonance imaging (Cine MRI) was performed in 20 patients aged 19 days to 13 years (mean 4.0 years), who had congenital heart disease confirmed at echocardiography or angiography. Prior to cine MRI, gated MRI was performed to evaluate for cardiac structure. Cine MRI was demonstrated by fast low fip angle shot imaging technique with a 30deg flip angle, 15 msec echo time, 30-40 msec pulse repetition time, and 128 x 128 acquisition matrix. Abnormalities of cardiac structure were extremely well defined in all patients by gated MRI. Intracardiac or intravascular blood flow were visualized in 17 (85%) of 20 patients by cine MRI. Left to right shunt flow through ventricular septal defect, atrial septal defect, and endocardial cushion defect were visualized with low signal intensity area. Low intensity jets flow through the site of re-coarctation of the aorta were also visualized. However, the good recording of cine MRI was not obtained because of artifacts in 3 of 20 patients (15%) who had severe congestive heart failure or respiratory arrhythmia. Gated MRI provides excellent visualization of fine structure, and cine MRI can provide high spatial resolution imaging of flow dynamic in a variety of congenital heart disease, noninvasively. (author)

  16. Cardiac cine imaging at 3 Tesla: initial experience with a 32-element body-array coil.

    Science.gov (United States)

    Fenchel, Michael; Deshpande, Vibhas S; Nael, Kambiz; Finn, J Paul; Miller, Stephan; Ruehm, Stefan; Laub, Gerhard

    2006-08-01

    We sought to assess the feasibility of cardiac cine imaging and evaluate image quality at 3 T using a body-array coil with 32 coil elements. Eight healthy volunteers (3 men; median age 29 years) were examined on a 3-T magnetic resonance scanner (Magnetom Trio, Siemens Medical Solutions) using a 32-element phased-array coil (prototype from In vivo Corp.). Gradient-recalled-echo (GRE) cine (GRAPPAx3), GRE cine with tagging lines, steady-state-free-precession (SSFP) cine (GRAPPAx3 and x4), and SSFP cine(TSENSEx4 andx6) images were acquired in short-axis and 4-chamber view. Reference images with identical scan parameters were acquired using the total-imaging-matrix (Tim) coil system with a total of 12 coil elements. Images were assessed by 2 observers in a consensus reading with regard to image quality, noise and presence of artifacts. Furthermore, signal-to-noise values were determined in phantom measurements. In phantom measurements signal-to-noise values were increased by 115-155% for the various cine sequences using the 32-element coil. Scoring of image quality yielded statistically significant increased image quality with the SSFP-GRAPPAx4, SSFP-TSENSEx4, and SSFP-TSENSEx6 sequence using the 32-element coil (P < 0.05). Similarly, scoring of image noise yielded a statistically significant lower noise rating with the SSFP-GRAPPAx4, GRE-GRAPPAx3, SSFP-TSENSEx4, and SSFP-TSENSEx6 sequence using the 32-element coil (P < 0.05). This study shows that cardiac cine imaging at 3 T using a 32-element body-array coil is feasible in healthy volunteers. Using a large number of coil elements with a favorable sensitivity profile supports faster image acquisition, with high diagnostic image quality even for high parallel imaging factors.

  17. Cardiovascular cine imaging and flow evaluation using Fast Interrupted Steady-State (FISS) magnetic resonance.

    Science.gov (United States)

    Edelman, Robert R; Serhal, Ali; Pursnani, Amit; Pang, Jianing; Koktzoglou, Ioannis

    2018-02-19

    Existing cine imaging techniques rely on balanced steady-state free precession (bSSFP) or spoiled gradient-echo readouts, each of which has limitations. For instance, with bSSFP, artifacts occur from rapid through-plane flow and off-resonance effects. We hypothesized that a prototype cine technique, radial fast interrupted steady-state (FISS), could overcome these limitations. The technique was compared with standard cine bSSFP for cardiac function, coronary artery conspicuity, and aortic valve morphology. Given its advantageous properties, we further hypothesized that the cine FISS technique, in combination with arterial spin labeling (ASL), could provide an alternative to phase contrast for visualizing in-plane flow patterns within the aorta and branch vessels. The study was IRB-approved and subjects provided consent. Breath-hold cine FISS and bSSFP were acquired using similar imaging parameters. There was no significant difference in biplane left ventricular ejection fraction or cardiac image quality between the two techniques. Compared with cine bSSFP, cine FISS demonstrated a marked decrease in fat signal which improved conspicuity of the coronary arteries, while suppression of through-plane flow artifact on thin-slice cine FISS images improved visualization of the aortic valve. Banding artifacts in the subcutaneous tissues were reduced. In healthy subjects, dynamic flow patterns were well visualized in the aorta, coronary and renal arteries using cine FISS ASL, even when the slice was substantially thicker than the vessel diameter. Cine FISS demonstrates several benefits for cardiovascular imaging compared with cine bSSFP, including better suppression of fat signal and reduced artifacts from through-plane flow and off-resonance effects. The main drawback is a slight (~ 20%) decrease in temporal resolution. In addition, preliminary results suggest that cine FISS ASL provides a potential alternative to phase contrast techniques for in-plane flow

  18. Myocardial strain assessment by cine cardiac magnetic resonance imaging using non-rigid registration.

    Science.gov (United States)

    Tsadok, Yossi; Friedman, Zvi; Haluska, Brian A; Hoffmann, Rainer; Adam, Dan

    2016-05-01

    To evaluate a novel post-processing method for assessment of longitudinal mid-myocardial strain in standard cine cardiac magnetic resonance (CMR) imaging sequences. Cine CMR imaging and tagged cardiac magnetic resonance imaging (TMRI) were performed in 15 patients with acute myocardial infarction (AMI) and 15 healthy volunteers served as control group. A second group of 37 post-AMI patients underwent both cine CMR and late gadolinium enhancement (LGE) CMR exams. Speckle tracking echocardiography (STE) was performed in 36 of these patients. Cine CMR, TMRI and STE were analyzed to obtain longitudinal strain. LGE-CMR datasets were analyzed to evaluate scar extent. Comparison of peak systolic strain (PSS) measured from CMR and TMRI yielded a strong correlation (r=0.86, pcine CMR data. The method was found to be highly correlated with strain measurements obtained by TMRI and STE. This tool allows accurate discrimination between different transmurality states of myocardial infarction. Copyright © 2015 Elsevier Inc. All rights reserved.

  19. A simple approach for EPID dosimetric calibration to overcome the effect of image-lag and ghosting

    International Nuclear Information System (INIS)

    Alshanqity, Mukhtar; Duane, Simon; Nisbet, Andrew

    2012-01-01

    EPID dosimetry has known drawbacks. The main issue is that a measurable residual signal is observed after the end of irradiation for prolonged periods of time, thus making measurement difficult. We present a detailed analysis of EPID response and suggest a simple, yet accurate approach for calibration that avoids the complexity of incorporating ghosting and image-lag using the maximum integrated signal instead of the total integrated signal. This approach is linear with dose and independent of dose rate. - Highlights: ► Image-lag and ghosting effects dosimetric accuracy. ► Image-lag and ghosting result in the reduction of total integrated signal for low doses. ► Residual signal is the most significant result for the image-lag and ghosting effects. ► Image-lag and ghosting can result in under-dosing of up to 2.5%.

  20. Usefulness of true FISP cine MR imaging in patients with poor cardiac function

    International Nuclear Information System (INIS)

    Sakuma, Toshiharu; Yamada, Naoaki; Motooka, Makoto; Enomoto, Naoyuki; Maeshima, Isamu; Matsuda, Kazuhide; Urayama, Shinichi; Ikeo, Miki

    2002-01-01

    This study was done to assess the value of True FISP cine in patients with poor cardiac function. True FISP cine and FLASH cine imaging were performed on a 1.5 T machine. Both short axis and horizontal long axis imaging sections were used. The imaging sections used a Matrix (120 x 128), FOV (24 x 32 cm), and had a slice thickness of 8 mm. The imaging time for True FISP cine was 8 heart beats and 17 heart beats for FLASH cine. The contrast-to-noise ratio between the blood and myocardium (CNR) was measured at enddiastole and endsystole. The subjects in the study were 10 healty volunteers (average age 26.5±3.2 years) and 12 patients with hypofunction (average age 53.9±13.2 years). In the volunteers, the CNR of the short axis imaging was similar in both True FISP (24.6±3.7) and FLASH (23.4±5.9). In the patients with poor cardiac function however, the CNR of True FISP was larger than FLASH in both the short and long axis. In the short axis (22.7±6.1 vs. 17.9±5.3, P<0.01) and in the long axis (17.4±4.3 vs. 9.3±4.0, P<0.01). We conclude that True FISP cine has a higher contrast in a shorter imaging time than FLASH cine. True FISP cine is especially useful in patients with poor cardiac function. (author)

  1. Motion-Corrected Real-Time Cine Magnetic Resonance Imaging of the Heart: Initial Clinical Experience.

    Science.gov (United States)

    Rahsepar, Amir Ali; Saybasili, Haris; Ghasemiesfe, Ahmadreza; Dolan, Ryan S; Shehata, Monda L; Botelho, Marcos P; Markl, Michael; Spottiswoode, Bruce; Collins, Jeremy D; Carr, James C

    2018-01-01

    Free-breathing real-time (RT) imaging can be used in patients with difficulty in breath-holding; however, RT cine imaging typically experiences poor image quality compared with segmented cine imaging because of low resolution. Here, we validate a novel unsupervised motion-corrected (MOCO) reconstruction technique for free-breathing RT cardiac images, called MOCO-RT. Motion-corrected RT uses elastic image registration to generate a single heartbeat of high-quality data from a free-breathing RT acquisition. Segmented balanced steady-state free precession (bSSFP) cine images and free-breathing RT images (Cartesian, TGRAPPA factor 4) were acquired with the same spatial/temporal resolution in 40 patients using clinical 1.5 T magnetic resonance scanners. The respiratory cycle was estimated using the reconstructed RT images, and nonrigid unsupervised motion correction was applied to eliminate breathing motion. Conventional segmented RT and MOCO-RT single-heartbeat cine images were analyzed to evaluate left ventricular (LV) function and volume measurements. Two radiologists scored images for overall image quality, artifact, noise, and wall motion abnormalities. Intraclass correlation coefficient was used to assess the reliability of MOCO-RT measurement. Intraclass correlation coefficient showed excellent reliability (intraclass correlation coefficient ≥ 0.95) of MOCO-RT with segmented cine in measuring LV function, mass, and volume. Comparison of the qualitative ratings indicated comparable image quality for MOCO-RT (4.80 ± 0.35) with segmented cine (4.45 ± 0.88, P = 0.215) and significantly higher than conventional RT techniques (3.51 ± 0.41, P cine (1.51 ± 0.90, P = 0.088 and 1.23 ± 0.45, P = 0.182) were not different. Wall motion abnormality ratings were comparable among different techniques (P = 0.96). The MOCO-RT technique can be used to process conventional free-breathing RT cine images and provides comparable quantitative assessment of LV function and volume

  2. Application of phase-contrast cine magnetic resonance imaging in endoscopic aqueductoplasty.

    Science.gov (United States)

    Chen, Guoqiang; Zheng, Jiaping; Xiao, Qing; Liu, Yunsheng

    2013-06-01

    The aim of this study was to evaluate the application of phase-contrast cine magnetic resonance imaging (MRI) in endoscopic aqueductoplasty (EA) for patients with obstructive hydrocephalus. The clinical diagnosis of hydrocephalus caused by aqueduct obstruction in 23 patients was confirmed by phase-contrast cine MRI examination. The patients were treated with EA and MRI was repeated during the follow-up. The cerebrospinal fluid (CSF) flow velocity in the aqueduct was measured to determine whether the aqueduct was obstructed. The results of phase-contrast cine MRI examinations indicated that there was no CSF flow in the aqueduct for all patients prior to surgery. Aqueductoplasty was successfully performed in all patients. The results of phase-contrast cine MRI examinations performed a week after surgery demonstrated an average CSF flow velocity of 4.74±1.77 cm/sec. During the follow-up, intracranial hypertension recurred in two patients in whom CSF flow was not observed in the aqueduct by the phase-contrast cine MRI scan. Aqueduct re-occlusion was revealed by an endoscopic exploration. By measuring the CSF flow velocity, phase-contrast cine MRI accurately identifies aqueduct obstruction. Cine MRI is a nontraumatic, simple and reliable method for determining whether the aqueduct is successfully opened following aqueductoplasty.

  3. Usefulness of breath-hold cardiac cine MR imaging with a middle field MRI system

    Energy Technology Data Exchange (ETDEWEB)

    Matsumura, Kentaro; Sato, Kiyoto; Aono, Masaki; Inoshita, Kenji; Utsumi, Naoko [Kagawa Inoshita Hospital, Ohnohara (Japan)

    1999-07-01

    To assess the accuracy of contrast-enhanced, single breath-hold cine MR imaging in calculating left ventricular volume and ejection fraction, we compared MR measurements with those obtained by using cine ventriculography in 60 patients. Fast cine MR images were acquired with a middle field MR system (0.5 T). A breath-hold single slice multi-phase fast gradient-echo (Fast Card) sequence was used to obtain fast cine MR images with the following parameters; TR of 16 ms, TE of 3 ms, flip angle of 30 degree, matrix elements of 256 x 128, view per segment of 6, field of view of 350 x 260 mm and one excitation. Left ventricular end-diastolic volume and ejection fraction obtained with contrast-enhanced Fast Card correlated well with those obtained with cine ventriculography (end-diastolic volume, y=1.00x+14.0, r=0.904, p<0.001; ejection fraction, y=0.961x+2.8, r=0.936, p<0.001). Our results show that contrast enhanced breath-hold cardiac cine MR imaging on horizontal long-axis view using a middle field MR system is an accurate method for evaluating left ventricular volume and ejection fraction. (author)

  4. Cine viability magnetic resonance imaging of the heart without increased scan time.

    Science.gov (United States)

    Hassanein, Azza S; Khalifa, Ayman M; Ibrahim, El-Sayed H

    2016-02-01

    Cardiac magnetic resonance imaging (MRI) provides information about myocardial morphology, function, and viability from cine, tagged, and late gadolinium enhancement (LGE) images, respectively. While the cine and tagged images are acquired in a time-resolved fashion, the LGE images are acquired at a single timeframe. The purpose of this work is to develop a method for generating cine LGE images without additional scan time. The motion field is extracted from the tagged images, and is then used to guide the deformation of the infarcted region from the acquired LGE image at the acquired timeframe to any other timeframe. Major techniques for motion estimation, including harmonic phase (HARP) and optical flow analysis, are tested in this work for motion estimation. The proposed method is tested on numerical phantom and images from four human subjects. The generated cine LGE images showed both viability and wall motion information in the same set of images without additional scan time or image misregistration problems. The band-pass optical flow analysis resulted in the most accurate motion estimation compared to other methods, especially HARP, which fails to track points at the myocardial boundary. Infarct transmurality from the generated images showed good agreement with myocardial strain, and wall thickening showed good agreement with that measured from conventional cine images. In conclusion, the developed technique allows for generating cine LGE images that enable simultaneous display of wall motion and viability information. The generated images could be useful for estimating myocardial contractility reserve and for treatment prognosis. Copyright © 2015 Elsevier Inc. All rights reserved.

  5. Overcoming pitfalls in the diagnosis of aortic dissection with phase-contrast cine MR imaging

    International Nuclear Information System (INIS)

    Herfkens, R.J.; Trefelner, E.C.; Jeffrey, R.B.; Pelc, N.J.; Steiner, R.M.; Francis, I.R.

    1991-01-01

    This paper evaluates whether phase contrast cine MR imaging improves the diagnostic accuracy in aortic dissection. Fifty-five MR examinations in 49 patients were analyzed. Gated coronal and axial spin-echo (SE) images and axial velocity encoded cine images were collected through the thoracic aorta. Blinded interpretations were obtained by first evaluating SE images, then SE and cine (magnitude) images, and finally with the addition of the velocity data for whether there was increase or decreased diagnostic confidence of aortic dissection. Comparisons were made with CT, angiography, and follow-up (including surgery). Fisher discriminant analysis showed that a statistically significant improvement in diagnostic accuracy was made by the addition of the velocity data in cases where a definite diagnosis of aortic dissection could not be made from the SE images alone

  6. Peristalsis gap sign at cine magnetic resonance imaging for diagnosing strangulated small bowel obstruction. Feasibility study

    International Nuclear Information System (INIS)

    Takahara, Taro; Kwee, T.C.; Haradome, Hiroki

    2011-01-01

    The aim of this study was to determine the feasibility of cine magnetic resonance imaging (MRI) for diagnosing strangulated small bowel obstruction (SBO). This study included 38 patients with clinically confirmed SBO who had undergone cine MRI. Cine MRI scans were evaluated regarding the presence of the 'peristalsis gap sign' (referring to an akinetic or severely hypokinetic closed loop), indicating strangulation. Computed tomography (CT) was performed in 34 of 38 patients with (n=25) or without (n=9) contrast enhancement. CT images were evaluated using a combination of criteria (presence of hyperattenuation, poor contrast enhancement, mesenteric edema, wall thickening, massive ascites) indicating strangulation. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of cine MRI and CT for the diagnosis of strangulation were calculated and compared using surgical findings and the clinical course as the reference standard. Sensitivity, specificity, PPV, and NPV of cine MRI were 100%, 92.9%, 83.3%, and 100%, respectively; and those of CT (of which 26.5% was performed without contrast enhancement) were 66.7%, 92.0%, 75.0%, and 88.5%, respectively. There was no significant difference in diagnostic accuracy between the two methods (P=0.375). Cine MRI is a feasible and promising technique for diagnosing strangulation. (author)

  7. Evaluation of atrial, ventricular and atrioventricular septal defects by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Akagi, Teiji; Kato, Hirohisa; Kiyomatsu, Yumi; Saiki, Kuninobu; Suzuki, Kazushige; Eto, Takaharu

    1992-01-01

    Cine magnetic resonance imaging (MRI) was performed on 20 patients (mean age: 5.3±4.4 years) with atrial, ventricular, or atrioventricular septal defects for evaluation of cardiac structure and blood flow. Prior to cine MRI, electrocardiographycally gated MRI using multislice scquisition was performed on all patients to localize optimal slice location. Cine-MRI was obtained with a 30 deg flip angle, 15 msec echo time, and 30 msec pulse repetition time, on a 256 x 256 or 128 x 128 acquisition matrix. Abnormalities of cardiac structure were well defined in all patients by gated cardiac imaging. In 18 of the 20 patients, cine-MRI was able to detect shunt flow, visualized as a low intensity signal in comparison with the surrounding blood flow. Cine-MRI can provide not only accurate anatomy of cardiac structures but functional assessment of the cardiac chamber, wall topology and flow relations. Cine-MRI will become an important noninvasive technique for assessment of anatomy and physiology in congenital heart disease. (author)

  8. SU-E-J-234: Application of a Breathing Motion Model to ViewRay Cine MR Images

    International Nuclear Information System (INIS)

    O’Connell, D. P.; Thomas, D. H.; Dou, T. H.; Lamb, J. M.; Yang, L.; Low, D. A.

    2015-01-01

    Purpose: A respiratory motion model previously used to generate breathing-gated CT images was used with cine MR images. Accuracy and predictive ability of the in-plane models were evaluated. Methods: Sagittalplane cine MR images of a patient undergoing treatment on a ViewRay MRI/radiotherapy system were acquired before and during treatment. Images were acquired at 4 frames/second with 3.5 × 3.5 mm resolution and a slice thickness of 5 mm. The first cine frame was deformably registered to following frames. Superior/inferior component of the tumor centroid position was used as a breathing surrogate. Deformation vectors and surrogate measurements were used to determine motion model parameters. Model error was evaluated and subsequent treatment cines were predicted from breathing surrogate data. A simulated CT cine was created by generating breathing-gated volumetric images at 0.25 second intervals along the measured breathing trace, selecting a sagittal slice and downsampling to the resolution of the MR cines. A motion model was built using the first half of the simulated cine data. Model accuracy and error in predicting the remaining frames of the cine were evaluated. Results: Mean difference between model predicted and deformably registered lung tissue positions for the 28 second preview MR cine acquired before treatment was 0.81 +/− 0.30 mm. The model was used to predict two minutes of the subsequent treatment cine with a mean accuracy of 1.59 +/− 0.63 mm. Conclusion: Inplane motion models were built using MR cine images and evaluated for accuracy and ability to predict future respiratory motion from breathing surrogate measurements. Examination of long term predictive ability is ongoing. The technique was applied to simulated CT cines for further validation, and the authors are currently investigating use of in-plane models to update pre-existing volumetric motion models used for generation of breathing-gated CT planning images

  9. Cine MR imaging of uterine peristalsis in patients with endometriosis

    International Nuclear Information System (INIS)

    Kido, Aki; Togashi, Kaori; Koyama, Takashi; Fujimoto, Ryota; Nishino, Mizuki; Miyake, Kanae; Hayakawa, Katsumi; Iwasaku, Kazuhiro; Fujii, Shingo

    2007-01-01

    Endometriosis is one of the most important causes of infertility; however the precise mechanism by which it affects female fertility is unclear. The objective of this study was to study the functional aspects of the uterus by evaluating uterine contractility in patients with endometrial cysts of the ovary. The study population was recruited from two institutes and consisted of 26 women (periovulatory (10), luteal (13), and menstrual phase (3); age range: 19-51 years) with untreated endometriosis; the control group consisted of 12 healthy women (age range: 22-41 years). Cine MR imaging obtained by a 1.5T magnet was visually evaluated at 12 x faster than real speed, focusing on the presence of uterine peristalsis, the direction and frequency of peristalsis, and the presence of sustained uterine contractions. Uterine peristalsis was identifiable in 3/10, 3/13, and 3/3 of the endometriosis patients in each menstrual cycle, respectively, and in 11/12, 3/12, and 5/12 of their control subjects. Peristaltic detection rate and frequency were significantly less for the endometriosis group than for the controls in the periovulatory phase only (p<0.05). Sustained contractions were recognized in 19/36 control subjects and in 13/26 endometriosis patients, but the difference was not significant. Uterine peristalsis appears to be suppressed during the periovulatory phase in patients with endometriosis, which may have an adverse effect on sperm transport. (orig.)

  10. Cine MR imaging of uterine peristalsis in patients with endometriosis

    Energy Technology Data Exchange (ETDEWEB)

    Kido, Aki; Togashi, Kaori; Koyama, Takashi; Fujimoto, Ryota [Kyoto University, Department of Diagnostic Imaging and Nuclear Medicine, Graduate School of Medicine, Kyoto (Japan); Nishino, Mizuki [Beth Israel Deaconess Medical Center, Department of Radiology, Kyoto (Japan); Miyake, Kanae; Hayakawa, Katsumi [Kyoto City Hospital, Department of Radiology, Kyoto (Japan); Iwasaku, Kazuhiro [Kyoto City Hospital, Department of Obstetrics and Gynecology, Kyoto (Japan); Fujii, Shingo [Kyoto University, Department of Gynecology and Obstetrics, Kyoto (Japan)

    2007-07-15

    Endometriosis is one of the most important causes of infertility; however the precise mechanism by which it affects female fertility is unclear. The objective of this study was to study the functional aspects of the uterus by evaluating uterine contractility in patients with endometrial cysts of the ovary. The study population was recruited from two institutes and consisted of 26 women (periovulatory (10), luteal (13), and menstrual phase (3); age range: 19-51 years) with untreated endometriosis; the control group consisted of 12 healthy women (age range: 22-41 years). Cine MR imaging obtained by a 1.5T magnet was visually evaluated at 12 x faster than real speed, focusing on the presence of uterine peristalsis, the direction and frequency of peristalsis, and the presence of sustained uterine contractions. Uterine peristalsis was identifiable in 3/10, 3/13, and 3/3 of the endometriosis patients in each menstrual cycle, respectively, and in 11/12, 3/12, and 5/12 of their control subjects. Peristaltic detection rate and frequency were significantly less for the endometriosis group than for the controls in the periovulatory phase only (p<0.05). Sustained contractions were recognized in 19/36 control subjects and in 13/26 endometriosis patients, but the difference was not significant. Uterine peristalsis appears to be suppressed during the periovulatory phase in patients with endometriosis, which may have an adverse effect on sperm transport. (orig.)

  11. Three-dimensional visualization of myocardial motion and blood flow with cine-MR images

    International Nuclear Information System (INIS)

    Oshiro, Osamu; Matani, Ayumu; Chihara, Kunihiro; Mikami, Taisei; Kitabatake, Akira.

    1997-01-01

    This paper describes a three-dimensional (3D) reconstruction and presentation method to visualize myocardial motion and blood flow in a heart using cine-MR (magnetic resonance) images. Firstly, the region of myocardium and blood were segmented with certain threshold gray values. Secondly, some slices were interpolated linearly to reconstruct a 3D static image. Finally, a 3D dynamic image was presented with displaying the 3D static images sequentially. The experimental results indicate that this method enables to visualize not only normal but also abnormal blood flow in cine-mode. (author)

  12. Subacute and Chronic Left Ventricular Myocardial Scar: Accuracy of Texture Analysis on Nonenhanced Cine MR Images.

    Science.gov (United States)

    Baessler, Bettina; Mannil, Manoj; Oebel, Sabrina; Maintz, David; Alkadhi, Hatem; Manka, Robert

    2018-01-01

    Purpose To test whether texture analysis (TA) allows for the diagnosis of subacute and chronic myocardial infarction (MI) on noncontrast material-enhanced cine cardiac magnetic resonance (MR) images. Materials and Methods In this retrospective, institutional review board-approved study, 120 patients who underwent cardiac MR imaging and showed large transmural (volume of enhancement on late gadolinium enhancement [LGE] images >20%, n = 72) or small (enhanced volume ≤20%, n = 48) subacute or chronic ischemic scars were included. Sixty patients with normal cardiac MR imaging findings served as control subjects. Regions of interest for TA encompassing the left ventricle were drawn by two blinded, independent readers on cine images in end systole by using a freely available software package. Stepwise dimension reduction and texture feature selection based on reproducibility, machine learning, and correlation analyses were performed for selecting features, enabling the diagnosis of MI on nonenhanced cine MR images by using LGE imaging as the standard of reference. Results Five independent texture features allowed for differentiation between ischemic scar and normal myocardium on cine MR images in both subgroups: Teta1, Perc.01, Variance, WavEnHH.s-3, and S(5,5)SumEntrp (in patients with large MI: all P values cine MR images, with an area under the curve of 0.93 and 0.92, respectively. Conclusion This proof-of-concept study indicates that TA of nonenhanced cine MR images allows for the diagnosis of subacute and chronic MI with high accuracy. © RSNA, 2017 Online supplemental material is available for this article.

  13. Cine MR imaging before and after realignment surgery for patellar maltracking - comparison with axial radiographs

    Energy Technology Data Exchange (ETDEWEB)

    Brossmann, J. [Klinik fuer Radiologische Diagnostik der Univ. Kiel (Germany); Muhle, C. [Klinik fuer Radiologische Diagnostik der Univ. Kiel (Germany); Buell, C.C. [Rheumaklinik, Bad Bramstedt (Germany); Zieplies, J. [Lubinus Klinik, Kiel (Germany); Melchert, U.H. [Klinik fuer Radiologische Diagnostik der Univ. Kiel (Germany); Brinkmann, G. [Klinik fuer Radiologische Diagnostik der Univ. Kiel (Germany); Schroeder, C. [Klinik fuer Radiologische Diagnostik der Univ. Kiel (Germany); Heller, M. [Klinik fuer Radiologische Diagnostik der Univ. Kiel (Germany)

    1995-04-01

    Comparison of motion-triggered cine magnetic resonance (MR) imaging and conventional radiographs for the assessment of operative results of patellar realignment. Fifteen patients with recurrent patellar dislocation or patellar subluxation were evaluated with conventional axial radiographs before and after realignment surgery by measuring the congruence angle (CA), lateral patellofemoral angle (LPFA), and lateral displacement (d). In eight patients the patellofemoral joint was additionally evaluated pre- and postoperatively with motion-triggered cine MR imaging by determining the bisect offset (BSO), lateral patellar displacement (LPD), and patellar tilt angle (PTA). Significant differences between the pre- and postoperative measurements were found for all MR imaging parameters (BSO, LPD, PTA: p<0.01) but not for the conventional X-ray parameters (CA: p=0.70, LPFA: p=0.56; d: p=0.04). Motion-triggered cine MR imaging was superior to conventional tangential radiographs for assessing the effectiveness of patellar realignment surgery. (orig.)

  14. Cine MR imaging before and after realignment surgery for patellar maltracking - comparison with axial radiographs

    International Nuclear Information System (INIS)

    Brossmann, J.; Muhle, C.; Buell, C.C.; Zieplies, J.; Melchert, U.H.; Brinkmann, G.; Schroeder, C.; Heller, M.

    1995-01-01

    Comparison of motion-triggered cine magnetic resonance (MR) imaging and conventional radiographs for the assessment of operative results of patellar realignment. Fifteen patients with recurrent patellar dislocation or patellar subluxation were evaluated with conventional axial radiographs before and after realignment surgery by measuring the congruence angle (CA), lateral patellofemoral angle (LPFA), and lateral displacement (d). In eight patients the patellofemoral joint was additionally evaluated pre- and postoperatively with motion-triggered cine MR imaging by determining the bisect offset (BSO), lateral patellar displacement (LPD), and patellar tilt angle (PTA). Significant differences between the pre- and postoperative measurements were found for all MR imaging parameters (BSO, LPD, PTA: p<0.01) but not for the conventional X-ray parameters (CA: p=0.70, LPFA: p=0.56; d: p=0.04). Motion-triggered cine MR imaging was superior to conventional tangential radiographs for assessing the effectiveness of patellar realignment surgery. (orig.)

  15. The Role of Cine Flow Magnetic Resonance Imaging in Patients with Chiari 0 Malformation.

    Science.gov (United States)

    Ozsoy, Kerem Mazhar; Oktay, Kadir; Cetinalp, Nuri Eralp; Gezercan, Yurdal; Erman, Tahsin

    2018-01-01

    To define the role of phase-contrast cine magnetic resonance imaging (MRI) in deciding the therapeutic strategy and underlying pathophysiology resulting in syrinx formation in patients with Chiari type 0 malformation. Seven patients who were admitted to our clinic with the diagnosis of Chiari 0 malformation from January 2005 to July 2016 were enrolled in the study. All patients underwent a detailed preoperative neurological examination. Entire neuroaxis MRI and phase-contrast cine MRI were obtained preoperatively and postoperatively. Seven patients (5 female and 2 male) with Chiari type 0 malformation fulfilled the inclusion criteria. All of the patients had absent cine flow at the craniovertebral junction except two patients. These five patients underwent surgical interventions; suboccipital decompression and duraplasty. All of them showed both clinical and radiological improvement in the postoperative period. Cine flow MRI appears to be a useful tool in the management of patients with Chiari 0 malformation. There was a good correlation between the clinical presentation and cine flow preoperatively, and between clinical improvement and cine flow in the postoperative period.

  16. Dosimetric characterization of an electronic portal imaging device (EPID) and development of a portal dosimetry simple model; Caracterizacion dosimetrica de un dispositivo electronico de imagen portal (EPID) y desarrollo de un modelo simple de dosimetria portal

    Energy Technology Data Exchange (ETDEWEB)

    Ripol ValentIn, O.; GarcIa Romero, A.; Hernandez Vitoria, A.; Jimenez Albericio, J.; Cortes Rodicio, J.; Millan Cebrian, E.; Ruiz Manzano, P.; Canellas Anoz, M.

    2010-07-01

    The use of the Electronic Portal Imaging Devices (EPID) for the quality control of linear accelerators of electrons is increasingly extended in practice. In this work the dosimetric characteristics of an EPID OptiVue{sup TM}1000 ST were studied and a friendly and simple method for the absorbed dose calibration was suggested. This method is based on a simple mathematical model, including: an absorbed dose transformation coefficient and image lag and field shape corrections. Software tools were developed in order to process the information and the results were validated by comparing them with the measured data with ionization chambers. The studied device showed suitable characteristics for its use for EPID dosimetry and the calculated results fitted satisfactorily with the dose planes obtained with the ionization chambers. Keeping in mind the model limitations, we concluded that it is possible to start the use of the EPID for the accelerator quality control and improvements for the current model should be studied, as well as other suitable applications: e.g. the Intensity Modulated Radiation Therapy (IMRT) treatment verification procedures. (Author).

  17. Free-breathing whole-heart 3D cine magnetic resonance imaging with prospective respiratory motion compensation.

    Science.gov (United States)

    Moghari, Mehdi H; Barthur, Ashita; Amaral, Maria E; Geva, Tal; Powell, Andrew J

    2018-07-01

    To develop and validate a new prospective respiratory motion compensation algorithm for free-breathing whole-heart 3D cine steady-state free precession (SSFP) imaging. In a 3D cine SSFP sequence, 4 excitations per cardiac cycle are re-purposed to prospectively track heart position. Specifically, their 1D image is reconstructed and routed into the scanner's standard diaphragmatic navigator processing system. If all 4 signals are in end-expiration, cine image data from the entire cardiac cycle is accepted for image reconstruction. Prospective validation was carried out in patients (N = 17) by comparing in each a conventional breath-hold 2D cine ventricular short-axis stack and a free-breathing whole-heart 3D cine data set. All 3D cine SSFP acquisitions were successful and the mean scan time was 5.9 ± 2.7 min. Left and right ventricular end-diastolic, end-systolic, and stroke volumes by 3D cine SSFP were all larger than those from 2D cine SSFP. This bias was 3D cine images had a lower ventricular blood-to-myocardium contrast ratio, contrast-to-noise ratio, mass, and subjective quality score. The novel prospective respiratory motion compensation method for 3D cine SSFP imaging was robust and efficient and yielded slightly larger ventricular volumes and lower mass compared to breath-hold 2D cine imaging. Magn Reson Med 80:181-189, 2018. © 2017 International Society for Magnetic Resonance in Medicine. © 2017 International Society for Magnetic Resonance in Medicine.

  18. SU-E-J-61: Monitoring Tumor Motion in Real-Time with EPID Imaging During Cervical Cancer Treatment

    International Nuclear Information System (INIS)

    Mao, W; Hrycushko, B; Yan, Y; Foster, R; Albuquerque, K

    2015-01-01

    Purpose: Traditional external beam radiotherapy for cervical cancer requires setup by external skin marks. In order to improve treatment accuracy and reduce planning margin for more conformal therapy, it is essential to monitor tumor positions interfractionally and intrafractionally. We demonstrate feasibility of monitoring cervical tumor motion online using EPID imaging from Beam’s Eye View. Methods: Prior to treatment, 1∼2 cylindrical radio opaque markers were implanted into inferior aspect of cervix tumor. During external beam treatments on a Varian 2100C by 4-field 3D plans, treatment beam images were acquired continuously by an EPID. A Matlab program was developed to locate internal markers on MV images. Based on 2D marker positions obtained from different treatment fields, their 3D positions were estimated for every treatment fraction. Results: There were 398 images acquired during different treatment fractions of three cervical cancer patients. Markers were successfully located on every frame of image at an analysis speed of about 1 second per frame. Intrafraction motions were evaluated by comparing marker positions relative to the position on the first frame of image. The maximum intrafraction motion of the markers was 1.6 mm. Interfraction motions were evaluated by comparing 3D marker positions at different treatment fractions. The maximum interfraction motion was up to 10 mm. Careful comparison found that this is due to patient positioning since the bony structures shifted with the markers. Conclusion: This method provides a cost-free and simple solution for online tumor tracking for cervical cancer treatment since it is feasible to acquire and export EPID images with fast analysis in real time. This method does not need any extra equipment or deliver extra dose to patients. The online tumor motion information will be very useful to reduce planning margins and improve treatment accuracy, which is particularly important for SBRT treatment with long

  19. SU-E-J-61: Monitoring Tumor Motion in Real-Time with EPID Imaging During Cervical Cancer Treatment

    Energy Technology Data Exchange (ETDEWEB)

    Mao, W; Hrycushko, B; Yan, Y; Foster, R; Albuquerque, K [UT Southwestern Medical Center, Dallas, TX (United States)

    2015-06-15

    Purpose: Traditional external beam radiotherapy for cervical cancer requires setup by external skin marks. In order to improve treatment accuracy and reduce planning margin for more conformal therapy, it is essential to monitor tumor positions interfractionally and intrafractionally. We demonstrate feasibility of monitoring cervical tumor motion online using EPID imaging from Beam’s Eye View. Methods: Prior to treatment, 1∼2 cylindrical radio opaque markers were implanted into inferior aspect of cervix tumor. During external beam treatments on a Varian 2100C by 4-field 3D plans, treatment beam images were acquired continuously by an EPID. A Matlab program was developed to locate internal markers on MV images. Based on 2D marker positions obtained from different treatment fields, their 3D positions were estimated for every treatment fraction. Results: There were 398 images acquired during different treatment fractions of three cervical cancer patients. Markers were successfully located on every frame of image at an analysis speed of about 1 second per frame. Intrafraction motions were evaluated by comparing marker positions relative to the position on the first frame of image. The maximum intrafraction motion of the markers was 1.6 mm. Interfraction motions were evaluated by comparing 3D marker positions at different treatment fractions. The maximum interfraction motion was up to 10 mm. Careful comparison found that this is due to patient positioning since the bony structures shifted with the markers. Conclusion: This method provides a cost-free and simple solution for online tumor tracking for cervical cancer treatment since it is feasible to acquire and export EPID images with fast analysis in real time. This method does not need any extra equipment or deliver extra dose to patients. The online tumor motion information will be very useful to reduce planning margins and improve treatment accuracy, which is particularly important for SBRT treatment with long

  20. Evaluation of valvular regurgitation by cine magnetic resonance imaging in patients with various cardiac diseases

    International Nuclear Information System (INIS)

    Kubota, Shuuhei; Nishimura, Tsunehiko

    1990-01-01

    In order to evaluate the clinical value and limitation of magnetic resonance imaging (MRI) for detection and quantification of valvular regurgitation, 98 patients with various cardiac diseases were studied by cine MRI and the results were compared with contrast angiography and doppler color-flow imaging. Cine MRI was carried out using FLASH (fast low angle shot) which employs TE of 10∼20 msec and TR of 30∼40 msec. 22 transverse tomograms per cardiac cycle with a slice thickness of 10 mm were obtained at the level of atrium and ventricle. The jet of valvular regurgitation was easily seen as a discrete are of low signal with cine MRI. Identification of the regurgitation and its severity were visually evaluated based on the relative size of the regurgitant jet from the incompetent valve orifice. Using contrast angiography as a gold standard, the sensitivity of cine MRI for detecting mitral regurgitation was 83% and was 94% for aortic regurgitation, with the specificity of 82% and 100%, respectively. For mitral requrgitation and aortic regurgitation, evaluation by cine MRI and severity agreed well with contrast angiography. By the comparative study with doppler color-flow imaging, relatively good agreement was found between the two methods in detection and quantitative evaluation of valvular regurgitation in any of four valves. Cine MRI was suggested to be useful for both the detection and semiquantification of valvular regurgitation in generally, but its clinical limitation at this point was also found because, 1)its images are not acquired in real times, as in contrast angiography or doppler color-flow imaging, but are compiled from the cumulative information from 128 heart beats, 2)the evaluation of regurgitation is made from only two-dimensional transverse tomograms. (author)

  1. Strain analysis in CRT candidates using the novel segment length in cine (SLICE) post-processing technique on standard CMR cine images

    International Nuclear Information System (INIS)

    Zweerink, Alwin; Allaart, Cornelis P.; Wu, LiNa; Beek, Aernout M.; Rossum, Albert C. van; Nijveldt, Robin; Kuijer, Joost P.A.; Ven, Peter M. van de; Meine, Mathias; Croisille, Pierre; Clarysse, Patrick

    2017-01-01

    Although myocardial strain analysis is a potential tool to improve patient selection for cardiac resynchronization therapy (CRT), there is currently no validated clinical approach to derive segmental strains. We evaluated the novel segment length in cine (SLICE) technique to derive segmental strains from standard cardiovascular MR (CMR) cine images in CRT candidates. Twenty-seven patients with left bundle branch block underwent CMR examination including cine imaging and myocardial tagging (CMR-TAG). SLICE was performed by measuring segment length between anatomical landmarks throughout all phases on short-axis cines. This measure of frame-to-frame segment length change was compared to CMR-TAG circumferential strain measurements. Subsequently, conventional markers of CRT response were calculated. Segmental strains showed good to excellent agreement between SLICE and CMR-TAG (septum strain, intraclass correlation coefficient (ICC) 0.76; lateral wall strain, ICC 0.66). Conventional markers of CRT response also showed close agreement between both methods (ICC 0.61-0.78). Reproducibility of SLICE was excellent for intra-observer testing (all ICC ≥0.76) and good for interobserver testing (all ICC ≥0.61). The novel SLICE post-processing technique on standard CMR cine images offers both accurate and robust segmental strain measures compared to the 'gold standard' CMR-TAG technique, and has the advantage of being widely available. (orig.)

  2. Strain analysis in CRT candidates using the novel segment length in cine (SLICE) post-processing technique on standard CMR cine images

    Energy Technology Data Exchange (ETDEWEB)

    Zweerink, Alwin; Allaart, Cornelis P.; Wu, LiNa; Beek, Aernout M.; Rossum, Albert C. van; Nijveldt, Robin [VU University Medical Center, Department of Cardiology, and Institute for Cardiovascular Research (ICaR-VU), Amsterdam (Netherlands); Kuijer, Joost P.A. [VU University Medical Center, Department of Physics and Medical Technology, Amsterdam (Netherlands); Ven, Peter M. van de [VU University Medical Center, Department of Epidemiology and Biostatistics, Amsterdam (Netherlands); Meine, Mathias [University Medical Center, Department of Cardiology, Utrecht (Netherlands); Croisille, Pierre; Clarysse, Patrick [Univ Lyon, UJM-Saint-Etienne, INSA, CNRS UMR 5520, INSERM U1206, CREATIS, Saint-Etienne (France)

    2017-12-15

    Although myocardial strain analysis is a potential tool to improve patient selection for cardiac resynchronization therapy (CRT), there is currently no validated clinical approach to derive segmental strains. We evaluated the novel segment length in cine (SLICE) technique to derive segmental strains from standard cardiovascular MR (CMR) cine images in CRT candidates. Twenty-seven patients with left bundle branch block underwent CMR examination including cine imaging and myocardial tagging (CMR-TAG). SLICE was performed by measuring segment length between anatomical landmarks throughout all phases on short-axis cines. This measure of frame-to-frame segment length change was compared to CMR-TAG circumferential strain measurements. Subsequently, conventional markers of CRT response were calculated. Segmental strains showed good to excellent agreement between SLICE and CMR-TAG (septum strain, intraclass correlation coefficient (ICC) 0.76; lateral wall strain, ICC 0.66). Conventional markers of CRT response also showed close agreement between both methods (ICC 0.61-0.78). Reproducibility of SLICE was excellent for intra-observer testing (all ICC ≥0.76) and good for interobserver testing (all ICC ≥0.61). The novel SLICE post-processing technique on standard CMR cine images offers both accurate and robust segmental strain measures compared to the 'gold standard' CMR-TAG technique, and has the advantage of being widely available. (orig.)

  3. Commissioning and quality assurance for VMAT delivery systems: An efficient time-resolved system using real-time EPID imaging.

    Science.gov (United States)

    Zwan, Benjamin J; Barnes, Michael P; Hindmarsh, Jonathan; Lim, Seng B; Lovelock, Dale M; Fuangrod, Todsaporn; O'Connor, Daryl J; Keall, Paul J; Greer, Peter B

    2017-08-01

    An ideal commissioning and quality assurance (QA) program for Volumetric Modulated Arc Therapy (VMAT) delivery systems should assess the performance of each individual dynamic component as a function of gantry angle. Procedures within such a program should also be time-efficient, independent of the delivery system and be sensitive to all types of errors. The purpose of this work is to develop a system for automated time-resolved commissioning and QA of VMAT control systems which meets these criteria. The procedures developed within this work rely solely on images obtained, using an electronic portal imaging device (EPID) without the presence of a phantom. During the delivery of specially designed VMAT test plans, EPID frames were acquired at 9.5 Hz, using a frame grabber. The set of test plans was developed to individually assess the performance of the dose delivery and multileaf collimator (MLC) control systems under varying levels of delivery complexities. An in-house software tool was developed to automatically extract features from the EPID images and evaluate the following characteristics as a function of gantry angle: dose delivery accuracy, dose rate constancy, beam profile constancy, gantry speed constancy, dynamic MLC positioning accuracy, MLC speed and acceleration constancy, and synchronization between gantry angle, MLC positioning and dose rate. Machine log files were also acquired during each delivery and subsequently compared to information extracted from EPID image frames. The largest difference between measured and planned dose at any gantry angle was 0.8% which correlated with rapid changes in dose rate and gantry speed. For all other test plans, the dose delivered was within 0.25% of the planned dose for all gantry angles. Profile constancy was not found to vary with gantry angle for tests where gantry speed and dose rate were constant, however, for tests with varying dose rate and gantry speed, segments with lower dose rate and higher gantry

  4. Higher resolution cine imaging with compressed sensing for accelerated clinical left ventricular evaluation.

    Science.gov (United States)

    Lin, Aaron C W; Strugnell, Wendy; Riley, Robyn; Schmitt, Benjamin; Zenge, Michael; Schmidt, Michaela; Morris, Norman R; Hamilton-Craig, Christian

    2017-06-01

    To assess the clinical feasibility of a compressed sensing cine magnetic resonance imaging (MRI) sequence of both high temporal and spatial resolution (CS_bSSFP) in comparison to a balanced steady-state free precession cine (bSSFP) sequence for reliable quantification of left ventricular (LV) volumes and mass. Segmented MRI cine images were acquired on a 1.5T scanner in 50 patients in the LV short-axis stack orientation using a retrospectively gated conventional bSSFP sequence (generalized autocalibrating partially parallel acquisition [GRAPPA] acceleration factor 2), followed by a prospectively triggered CS_bSSFP sequence with net acceleration factor of 8. Image quality was assessed by published criteria. Comparison of sequences was made in LV volumes and mass, image quality score, quantitative regional myocardial wall motion, and imaging time using Pearson's correlation, Bland-Altman and paired 2-tailed Student's t-test. Differences (bSSFP minus CS_bSSFP, mean ± SD) and Pearson's correlations were 14.8 ± 16.3 (P = 0.31) and r = 0.98 (P cine CS_bSSFP accurately and reliably quantitates LV volumes and mass, shortens acquisition times, and is clinically feasible. 1 Technical Efficacy: Stage 2 J. MAGN. RESON. IMAGING 2017;45:1693-1699. © 2016 International Society for Magnetic Resonance in Medicine.

  5. Accelerated two-dimensional cine DENSE cardiovascular magnetic resonance using compressed sensing and parallel imaging.

    Science.gov (United States)

    Chen, Xiao; Yang, Yang; Cai, Xiaoying; Auger, Daniel A; Meyer, Craig H; Salerno, Michael; Epstein, Frederick H

    2016-06-14

    Cine Displacement Encoding with Stimulated Echoes (DENSE) provides accurate quantitative imaging of cardiac mechanics with rapid displacement and strain analysis; however, image acquisition times are relatively long. Compressed sensing (CS) with parallel imaging (PI) can generally provide high-quality images recovered from data sampled below the Nyquist rate. The purposes of the present study were to develop CS-PI-accelerated acquisition and reconstruction methods for cine DENSE, to assess their accuracy for cardiac imaging using retrospective undersampling, and to demonstrate their feasibility for prospectively-accelerated 2D cine DENSE imaging in a single breathhold. An accelerated cine DENSE sequence with variable-density spiral k-space sampling and golden angle rotations through time was implemented. A CS method, Block LOw-rank Sparsity with Motion-guidance (BLOSM), was combined with sensitivity encoding (SENSE) for the reconstruction of under-sampled multi-coil spiral data. Seven healthy volunteers and 7 patients underwent 2D cine DENSE imaging with fully-sampled acquisitions (14-26 heartbeats in duration) and with prospectively rate-2 and rate-4 accelerated acquisitions (14 and 8 heartbeats in duration). Retrospectively- and prospectively-accelerated data were reconstructed using BLOSM-SENSE and SENSE. Image quality of retrospectively-undersampled data was quantified using the relative root mean square error (rRMSE). Myocardial displacement and circumferential strain were computed for functional assessment, and linear correlation and Bland-Altman analyses were used to compare accelerated acquisitions to fully-sampled reference datasets. For retrospectively-undersampled data, BLOSM-SENSE provided similar or lower rRMSE at rate-2 and lower rRMSE at rate-4 acceleration compared to SENSE (p cine DENSE provided good image quality and expected values of displacement and strain. BLOSM-SENSE-accelerated spiral cine DENSE imaging with 2D displacement encoding can be

  6. Automatic segmentation of left ventricle in cardiac cine MRI images based on deep learning

    Science.gov (United States)

    Zhou, Tian; Icke, Ilknur; Dogdas, Belma; Parimal, Sarayu; Sampath, Smita; Forbes, Joseph; Bagchi, Ansuman; Chin, Chih-Liang; Chen, Antong

    2017-02-01

    In developing treatment of cardiovascular diseases, short axis cine MRI has been used as a standard technique for understanding the global structural and functional characteristics of the heart, e.g. ventricle dimensions, stroke volume and ejection fraction. To conduct an accurate assessment, heart structures need to be segmented from the cine MRI images with high precision, which could be a laborious task when performed manually. Herein a fully automatic framework is proposed for the segmentation of the left ventricle from the slices of short axis cine MRI scans of porcine subjects using a deep learning approach. For training the deep learning models, which generally requires a large set of data, a public database of human cine MRI scans is used. Experiments on the 3150 cine slices of 7 porcine subjects have shown that when comparing the automatic and manual segmentations the mean slice-wise Dice coefficient is about 0.930, the point-to-curve error is 1.07 mm, and the mean slice-wise Hausdorff distance is around 3.70 mm, which demonstrates the accuracy and robustness of the proposed inter-species translational approach.

  7. Strain analysis in CRT candidates using the novel segment length in cine (SLICE) post-processing technique on standard CMR cine images.

    Science.gov (United States)

    Zweerink, Alwin; Allaart, Cornelis P; Kuijer, Joost P A; Wu, LiNa; Beek, Aernout M; van de Ven, Peter M; Meine, Mathias; Croisille, Pierre; Clarysse, Patrick; van Rossum, Albert C; Nijveldt, Robin

    2017-12-01

    Although myocardial strain analysis is a potential tool to improve patient selection for cardiac resynchronization therapy (CRT), there is currently no validated clinical approach to derive segmental strains. We evaluated the novel segment length in cine (SLICE) technique to derive segmental strains from standard cardiovascular MR (CMR) cine images in CRT candidates. Twenty-seven patients with left bundle branch block underwent CMR examination including cine imaging and myocardial tagging (CMR-TAG). SLICE was performed by measuring segment length between anatomical landmarks throughout all phases on short-axis cines. This measure of frame-to-frame segment length change was compared to CMR-TAG circumferential strain measurements. Subsequently, conventional markers of CRT response were calculated. Segmental strains showed good to excellent agreement between SLICE and CMR-TAG (septum strain, intraclass correlation coefficient (ICC) 0.76; lateral wall strain, ICC 0.66). Conventional markers of CRT response also showed close agreement between both methods (ICC 0.61-0.78). Reproducibility of SLICE was excellent for intra-observer testing (all ICC ≥0.76) and good for interobserver testing (all ICC ≥0.61). The novel SLICE post-processing technique on standard CMR cine images offers both accurate and robust segmental strain measures compared to the 'gold standard' CMR-TAG technique, and has the advantage of being widely available. • Myocardial strain analysis could potentially improve patient selection for CRT. • Currently a well validated clinical approach to derive segmental strains is lacking. • The novel SLICE technique derives segmental strains from standard CMR cine images. • SLICE-derived strain markers of CRT response showed close agreement with CMR-TAG. • Future studies will focus on the prognostic value of SLICE in CRT candidates.

  8. SU-C-BRD-06: Sensitivity Study of An Automated System to Acquire and Analyze EPID Exit Dose Images

    Energy Technology Data Exchange (ETDEWEB)

    Olch, A; Zhuang, A [University of Southern California, Los Angeles, CA (United States)

    2015-06-15

    Purpose: The dosimetric consequences of errors in patient setup or beam delivery and anatomical changes are not readily known. A new product, PerFRACTION (Sun Nuclear Corporation), is designed to detect these errors by comparing the EPID exit dose image from each field of each fraction to those from baseline fraction images. This work investigates the sensitivity of PerFRACTION to detect the deviation of induced errors in a variety of realistic scenarios. Methods: Eight plans were created mimicking potential delivery or setup errors. The plans consisted of a nominal field and the field with an induced error. These were used to irradiate the EPID simulating multiple fractions with and without the error. Integrated EPID images were acquired in clinical mode and saved in ARIA. PerFRACTION automatically pulls the images into its database and performs the user defined comparison. In some cases, images were manually pushed to PerFRACTION. We varied the distance-to-agreement or dose tolerance until PerFRACTION showed failing pixels in the affected region and recorded the values. We induced errors of 1mm and greater in jaw, MLC, and couch position, 2 degree collimation rotation (patient yaw), and 0.5% to 1.5% in machine output. Both static and arc fields with the rails in or out were also acquired and compared. Results: PerFRACTION detected position errors of the jaws, MLC, and couch with an accuracy of better than 0.5 mm, and 0.2 degrees for collimator and gantry error. PerFRACTION detected a machine output error within 0.2% and detected the change in rail position. Conclusion: A new automated system for monitoring daily treatments for machine or patient variations from the first fraction using integrated EPID images was found to be sensitive enough to detect small positional, angular, and dosimetric errors within 0.5mm, 0.2 degrees, and 0.2%, respectively. Sun Nuclear Corporation has provided a software license for the product described.

  9. Evaluation of esophageal peristalsis in patients with esophageal tumors. Initial experience with cine MR imaging

    International Nuclear Information System (INIS)

    Koyama, Takashi; Kobayashi, Ari; Hiraga, Akira; Umeoka, Shigeaki; Saga, Tsuneo; Watanabe, Go; Tamai, Ken; Shimada, Yutaka; Togashi, Kaori

    2005-01-01

    We evaluated esophageal peristalsis in patients with esophageal tumors by cine MR using steady-state free precession (SSFP) sequence and correlated the alteration of the esophageal peristalsis with clinical symptoms and tumor stages. Thirteen patients with pathologically proven esophageal tumors, including 12 esophageal cancers and one submucosal leiomyoma, underwent cine MRI using true fast imaging with steady precession (trueFISP) sequence, which is one SSFP sequence, after contrast-enhanced MR scanning for clinical purposes. A total of 120 serial images were obtained within 60 s through the plane along the long axis of the esophagus while patients chewed gum. The serial trueFISP images were evaluated for the presence, frequency, speed of progression, and passage of peristalsis through the tumor. The data from cine MRI were compared with clinical symptoms and tumor stages. Peristalsis was clearly identified in all patients. Seven patients with complete interruption of peristalsis had dysphagia; one with partially impaired peristalsis could intake solid foods with discomfort; and two with partially impaired peristalsis and three with preserved peristalsis remained asymptomatic. Patients with complete or partial interruption of peristalsis had Stage T3 or T4 esophageal cancer. In conclusion, trueFISP cine MR imaging enables direct visualization of esophageal peristalsis in relation to esophageal tumors. Complete interruption of peristalsis causes dysphagia, whereas partial interruption of and preserved peristalsis usually do not cause digestive problems. Interruption of peristalsis may indicate impaired muscle function caused by invasion of advanced esophageal cancers. (author)

  10. Functional imaging of murine hearts using accelerated self-gated UTE cine MRI

    NARCIS (Netherlands)

    Motaal, Abdallah G.; Noorman, Nils; de Graaf, Wolter L.; Hoerr, Verena; Florack, Luc M. J.; Nicolay, Klaas; Strijkers, Gustav J.

    2015-01-01

    We introduce a fast protocol for ultra-short echo time (UTE) Cine magnetic resonance imaging (MRI) of the beating murine heart. The sequence involves a self-gated UTE with golden-angle radial acquisition and compressed sensing reconstruction. The self-gated acquisition is performed asynchronously

  11. Analysis of 3-D Tongue Motion from Tagged and Cine Magnetic Resonance Images

    Science.gov (United States)

    Xing, Fangxu; Woo, Jonghye; Lee, Junghoon; Murano, Emi Z.; Stone, Maureen; Prince, Jerry L.

    2016-01-01

    Purpose: Measuring tongue deformation and internal muscle motion during speech has been a challenging task because the tongue deforms in 3 dimensions, contains interdigitated muscles, and is largely hidden within the vocal tract. In this article, a new method is proposed to analyze tagged and cine magnetic resonance images of the tongue during…

  12. Assessment of Bladder Motion for Clinical Radiotherapy Practice Using Cine-Magnetic Resonance Imaging

    International Nuclear Information System (INIS)

    McBain, Catherine A.; Khoo, Vincent S.; Buckley, David L.; Sykes, Jonathan S.; Green, Melanie M.; Cowan, Richard A.; Hutchinson, Charles E.; Moore, Christopher J.; Price, Patricia M.

    2009-01-01

    Purpose: Organ motion is recognized as the principal source of inaccuracy in bladder radiotherapy (RT), but there is currently little information on intrafraction bladder motion. Methods and Materials: We used cine-magnetic resonance imaging (cine-MRI) to study bladder motion relevant to intrafraction RT delivery. On two occasions, a 28 minute cine-MRI sequence was acquired from 10 bladder cancer patients and 5 control participants immediately after bladder emptying, after abstinence from drinking for the preceding hour. From the resulting cine sequences, bladder motion was subjectively assessed. To quantify bladder motion, the bladder was contoured in imaging volume sets at 0, 14, and 28 min to measure changes to bladder volumes, wall displacements, and center of gravity (COG) over time. Results: The dominant source of bladder motion during imaging was bladder filling (up to 101% volume increase); rectal and small bowel movements were transient, with minimal impact. Bladder volume changes were similar for all participants. However for bladder cancer patients, wall displacements were larger (up to 58 mm), less symmetrical, and more variable compared with nondiseased control bladders. Conclusions: Significant and individualized intrafraction bladder wall displacements may occur during bladder RT delivery. This important source of inaccuracy should be incorporated into treatment planning and verification.

  13. Functional imaging of murine hearts using accelerated self-gated UTE cine MRI.

    Science.gov (United States)

    Motaal, Abdallah G; Noorman, Nils; de Graaf, Wolter L; Hoerr, Verena; Florack, Luc M J; Nicolay, Klaas; Strijkers, Gustav J

    2015-01-01

    We introduce a fast protocol for ultra-short echo time (UTE) Cine magnetic resonance imaging (MRI) of the beating murine heart. The sequence involves a self-gated UTE with golden-angle radial acquisition and compressed sensing reconstruction. The self-gated acquisition is performed asynchronously with the heartbeat, resulting in a randomly undersampled kt-space that facilitates compressed sensing reconstruction. The sequence was tested in 4 healthy rats and 4 rats with chronic myocardial infarction, approximately 2 months after surgery. As a control, a non-accelerated self-gated multi-slice FLASH sequence with an echo time (TE) of 2.76 ms, 4.5 signal averages, a matrix of 192 × 192, and an acquisition time of 2 min 34 s per slice was used to obtain Cine MRI with 15 frames per heartbeat. Non-accelerated UTE MRI was performed with TE = 0.29 ms, a reconstruction matrix of 192 × 192, and an acquisition time of 3 min 47 s per slice for 3.5 averages. Accelerated imaging with 2×, 4× and 5× undersampled kt-space data was performed with 1 min, 30 and 15 s acquisitions, respectively. UTE Cine images up to 5× undersampled kt-space data could be successfully reconstructed using a compressed sensing algorithm. In contrast to the FLASH Cine images, flow artifacts in the UTE images were nearly absent due to the short echo time, simplifying segmentation of the left ventricular (LV) lumen. LV functional parameters derived from the control and the accelerated Cine movies were statistically identical.

  14. SU-D-201-06: Remote Dosmetric Auditing of VMAT Deliveries for Clinical Trials Using EPID

    Energy Technology Data Exchange (ETDEWEB)

    Legge, K; Miri, N [University of Newcastle (Australia); Lehmann, J [Calvary Mater Newcastle (Australia); Vial, P [Liverpool Hospital (Australia); Greer, P [University of Newcastle (Australia); Calvary Mater Newcastle (Australia)

    2016-06-15

    Purpose: To develop a method for remote dosimetric auditing the delivery of VMAT using EPID which allows for simple, inexpensive and time efficient dosimetric credentialing for clinical trials. Methods: Remote centers are provided with CT datasets and planning guidelines to produce VMAT plans for a head and neck and a post-prostatectomy treatment. Plans are transferred in the planning system to two virtual water equivalent phantoms, one flat and one cylindrical. Cine images are acquired during VMAT delivery to the EPID in air with gantry angle recorded in image headers. Centers also deliver provided calibration plans to enable EPID signal to dose conversion, determination of the central axis, and correction of EPID sag prior to analysis. EPID images and planned doses are sent to the central site. EPID cine images are converted to dose in the virtual phantoms using an established backprojection method (King et al., Med.Phys. 2012) with EPID backscatter correction. Individual arcs (with gantry angles collapsed to zero) are evaluated at 10 cm depth in the flat phantom using 2D gamma, and total doses are evaluated in the cylindrical phantom using 3D gamma. Results are reported for criteria of 3%,3mm, 3%,2mm and 2%,2mm for all points greater than 10% of global maximum. Results: The pilot study for Varian centers has commenced, and three centers have been audited for head and neck plans and two for post-prostatectomy plans to date. The mean pass rate for arc-by-arc 2D analysis at 3%,3mm is 99.5% and for 3D analysis is 95.8%. A method for Elekta linacs using an inclinometer for gantry angle information is under development. Conclusion: Preliminary results for this new method are promising. The method takes advantage of EPID equipment available at most centers and clinically established software to provide a feasible, low cost solution to credentialing centers for clinical trials. Funding has been provided from Calvary Mater Newcastle Department of Radiation Oncology, TROG

  15. SU-E-J-27: Shifting Multiple EPID Imager Layers to Improve Image Quality and Resolution in MV CBCT

    Energy Technology Data Exchange (ETDEWEB)

    Chen, H; Rottmann, J; Yip, S; Berbeco, R [Brigham and Women’s Hospital, Boston, Massachusetts (United States); Morf, D; Fueglistaller, R; Star-Lack, J; Zentai, G [Varian Medical Systems, Palo Alto, CA (United States)

    2015-06-15

    Purpose: Vertical stacking of four conventional EPID layers can improve DQE for MV-CBCT applications. We hypothesize that shifting each layer laterally by half a pixel relative to the layer above, will improve the contrast-to-noise ratio (CNR) and image resolution. Methods: For CNR assessment, a 20 cm diameter digital phantom with 8 inserts is created. The attenuation coefficient of the phantom is similar to lung at the average energy of a 6 MV photon beam. The inserts have attenuations 1, 2…8 times of lung. One of the inserts is close to soft tissue, resembling the case of a tumor in lung. For resolution assessment, a digital phantom featuring a bar pattern is created. The phantom has an attenuation coefficient similar to soft tissue and the bars have an attenuation coefficient of calcium sulfate. A 2 MeV photon beam is attenuated through these phantoms and hits each of the four stacked detector layers. Each successive layer is shifted by half a pixel in the x only, y only, and x and y (combined) directions, respectively. Blurring and statistical noise are added to the projections. Projections from one, two, three and four layers are used for reconstruction. CNR and image resolution are evaluated and compared. Results: When projections from multiple layers are combined for reconstruction, CNR increases with the number of layers involved. CNR in reconstructions from two, three and four layers are 1.4, 1.7 and 1.99 times that from one layer. The resolution from the shifted four layer detector is also improved from a single layer. In a comparison between one layer versus four layers in this preliminary study, the resolution from four shifted layers is at least 20% better. Conclusion: Layer-shifting in a stacked EPID imager design enhances resolution as well as CNR for half scan MV-CBCT. The project described was supported, in part, by a grant from Varian Medical Systems, Inc., and Award No. R01CA188446-01 from the National Cancer Institute. The content is solely

  16. Clinical usefulness of cardiac cine magnetic resonance imaging in patients with atrial fibrillation

    Energy Technology Data Exchange (ETDEWEB)

    Sakakura, Kazuyoshi; Anno, Hirofumi; Kondo, Takeshi [Fujita Health Univ., Toyoake, Aichi (Japan); and others

    1993-05-01

    We studied the clinical usefulness of cine mode magnetic resonance (MR) imaging in patients with atrial fibrillation (AF) from aspects of image quality and cardiac function. The signal-to-noise (S/N) ratio in the myocardium was significantly (p<0.05) lower in patients with AF than those with normal sinus rhythm. Two radiologists who did not know any patient's information evaluated the image quality visually by marking method on a scale of 12 points. There was no difference of image quality between the two groups. The standard deviation of R-R interval was significantly (r=-0.92, p<0.001) correlated with the S/N ratio in myocardium. Consequently, it was not favorable to estimate visually cardiac cine MR image in patients with AF, when standard deviation of R-R interval was large. The left ventricular (LV) end diastolic, end systolic and stroke volumes and ejection fraction were closely (r=0.82[approx]0.95, p<0.05[approx]0.001) correlated between MR imaging and M-mode echocardiography, respectively. The ability to detect left side valvular regurgitation was almost equal in both MR imaging and color Doppler echocardiography. This result was coincided to previous papers in patients with normal sinus rhythm. In conclusion, cine mode MR imaging was also useful to analyze cardiac function and detect valvular regurgitation in patients with AF. (author).

  17. Real-time SPARSE-SENSE cardiac cine MR imaging: optimization of image reconstruction and sequence validation.

    Science.gov (United States)

    Goebel, Juliane; Nensa, Felix; Bomas, Bettina; Schemuth, Haemi P; Maderwald, Stefan; Gratz, Marcel; Quick, Harald H; Schlosser, Thomas; Nassenstein, Kai

    2016-12-01

    Improved real-time cardiac magnetic resonance (CMR) sequences have currently been introduced, but so far only limited practical experience exists. This study aimed at image reconstruction optimization and clinical validation of a new highly accelerated real-time cine SPARSE-SENSE sequence. Left ventricular (LV) short-axis stacks of a real-time free-breathing SPARSE-SENSE sequence with high spatiotemporal resolution and of a standard segmented cine SSFP sequence were acquired at 1.5 T in 11 volunteers and 15 patients. To determine the optimal iterations, all volunteers' SPARSE-SENSE images were reconstructed using 10-200 iterations, and contrast ratios, image entropies, and reconstruction times were assessed. Subsequently, the patients' SPARSE-SENSE images were reconstructed with the clinically optimal iterations. LV volumetric values were evaluated and compared between both sequences. Sufficient image quality and acceptable reconstruction times were achieved when using 80 iterations. Bland-Altman plots and Passing-Bablok regression showed good agreement for all volumetric parameters. 80 iterations are recommended for iterative SPARSE-SENSE image reconstruction in clinical routine. Real-time cine SPARSE-SENSE yielded comparable volumetric results as the current standard SSFP sequence. Due to its intrinsic low image acquisition times, real-time cine SPARSE-SENSE imaging with iterative image reconstruction seems to be an attractive alternative for LV function analysis. • A highly accelerated real-time CMR sequence using SPARSE-SENSE was evaluated. • SPARSE-SENSE allows free breathing in real-time cardiac cine imaging. • For clinically optimal SPARSE-SENSE image reconstruction, 80 iterations are recommended. • Real-time SPARSE-SENSE imaging yielded comparable volumetric results as the reference SSFP sequence. • The fast SPARSE-SENSE sequence is an attractive alternative to standard SSFP sequences.

  18. SU-E-T-775: Use of Electronic Portal Imaging Device (EPID) for Quality Assurance (QA) of Electron Beams On Varian Truebeam System

    Energy Technology Data Exchange (ETDEWEB)

    Cai, B; Yaddanapudi, S; Sun, B; Li, H; Noel, C; Mutic, S; Goddu, S [Department of Radiation Oncology, Washington University in St Louis, St. Louis, MO (United States)

    2015-06-15

    Purpose: In a previous study we have demonstrated the feasibility of using EPID to QA electron beam parameters on a single Varian TrueBeam LINAC. This study aims to provide further investigation on (1) reproducibility of using EPID to detect electron beam energy changes on multiple machines and (2) evaluation of appropriate calibration methods to compare results from different EPIDs. Methods: Ad-hoc mode electron beam images were acquired in developer mode with XML code. Electron beam data were collected on a total of six machines from four institutions. A custom-designed double-wedge phantom was placed on the EPID detector. Two calibration methods - Pixel Sensitivity Map (PSM) and Large Source-to-Imager Distance Flood Field (LSID-FF) - were used. To test the sensitivity of EPID in detecting energy drifts, Bending Magnet Current (BMC) was detuned to invoke energy changes corresponding to ∼±1.5 mm change in R50% of PDD on two machines from two institutions. Percent depth ionization (PDI) curves were then analyzed and compared with the respective baseline images using LSID-FF calibration. For reproducibility testing, open field EPID images and images with a standard testing phantom were collected on multiple machines. Images with and without PSM correction for same energies on different machines were overlaid and compared. Results: Two pixel shifts were observed in PDI curve when energy changes exceeded the TG142 tolerance. PSM showed the potential to correct the differences in pixel response of different imagers. With PSM correction, the histogram of images differences obtained from different machines showed narrower distributions than those images without PSM correction. Conclusion: EPID is sensitive for electron energy changes and the results are reproducible on different machines. When overlaying images from different machines, PSM showed the ability to partially eliminate the intrinsic variation of various imagers. Research Funding from Varian Medical Systems

  19. TH-E-17A-10: Markerless Lung Tumor Tracking Based On Beams Eye View EPID Images

    Energy Technology Data Exchange (ETDEWEB)

    Chiu, T; Kearney, V; Liu, H; Jiang, L; Foster, R; Mao, W [UT Southwestern Medical Center, Dallas, Texas (United States); Rozario, T; Bereg, S [University of Texas at Dallas, Richardson, Texas (United States); Klash, S [Premier Cancer Centers, Dallas, TX (United States)

    2014-06-15

    Purpose: Dynamic tumor tracking or motion compensation techniques have proposed to modify beam delivery following lung tumor motion on the flight. Conventional treatment plan QA could be performed in advance since every delivery may be different. Markerless lung tumor tracking using beams eye view EPID images provides a best treatment evaluation mechanism. The purpose of this study is to improve the accuracy of the online markerless lung tumor motion tracking method. Methods: The lung tumor could be located on every frame of MV images during radiation therapy treatment by comparing with corresponding digitally reconstructed radiograph (DRR). A kV-MV CT corresponding curve is applied on planning kV CT to generate MV CT images for patients in order to enhance the similarity between DRRs and MV treatment images. This kV-MV CT corresponding curve was obtained by scanning a same CT electron density phantom by a kV CT scanner and MV scanner (Tomotherapy) or MV CBCT. Two sets of MV DRRs were then generated for tumor and anatomy without tumor as the references to tracking the tumor on beams eye view EPID images. Results: Phantom studies were performed on a Varian TrueBeam linac. MV treatment images were acquired continuously during each treatment beam delivery at 12 gantry angles by iTools. Markerless tumor tracking was applied with DRRs generated from simulated MVCT. Tumors were tracked on every frame of images and compared with expected positions based on programed phantom motion. It was found that the average tracking error were 2.3 mm. Conclusion: This algorithm is capable of detecting lung tumors at complicated environment without implanting markers. It should be noted that the CT data has a slice thickness of 3 mm. This shows the statistical accuracy is better than the spatial accuracy. This project has been supported by a Varian Research Grant.

  20. Imaging Three-Dimensional Myocardial Mechanics Using Navigator-gated Volumetric Spiral Cine DENSE MRI

    Science.gov (United States)

    Zhong, Xiaodong; Spottiswoode, Bruce S.; Meyer, Craig H.; Kramer, Christopher M.; Epstein, Frederick H.

    2010-01-01

    A navigator-gated 3D spiral cine displacement encoding with stimulated echoes (DENSE) pulse sequence for imaging 3D myocardial mechanics was developed. In addition, previously-described 2D post-processing algorithms including phase unwrapping, tissue tracking, and strain tensor calculation for the left ventricle (LV) were extended to 3D. These 3D methods were evaluated in 5 healthy volunteers, using 2D cine DENSE and historical 3D myocardial tagging as reference standards. With an average scan time of 20.5 ± 5.7 minutes, 3D data sets with a matrix size of 128 × 128 × 22, voxel size of 2.8 × 2.8 × 5.0 mm3, and temporal resolution of 32 ms were obtained with displacement encoding in three orthogonal directions. Mean values for end-systolic mid-ventricular mid-wall radial, circumferential, and longitudinal strain were 0.33 ± 0.10, −0.17 ± 0.02, and −0.16 ± 0.02, respectively. Transmural strain gradients were detected in the radial and circumferential directions, reflecting high spatial resolution. Good agreement by linear correlation and Bland-Altman analysis was achieved when comparing normal strains measured by 2D and 3D cine DENSE. Also, the 3D strains, twist, and torsion results obtained by 3D cine DENSE were in good agreement with historical values measured by 3D myocardial tagging. PMID:20574967

  1. Cine MR imaging in mitral valve prolapse; Study on mitral regurgitation and left atrial volume

    Energy Technology Data Exchange (ETDEWEB)

    Kumai, Toshihiko [Chiba Univ. (Japan). School of Medicine

    1993-02-01

    This study was undertaken to assess the ability of cine MR imaging to evaluate the direction, timing, and severity of mitral regurgitation in patients with mitral valve prolapse (MVP). The population of this study was 33 patients with MVP diagnosed by two-dimensional echocardiography and 10 patients with rheumatic mitral valve disease (MSR) for comparison. 7 patients with MVP and 5 with MSR had atrial fibrillation and/or history of congestive heart failure as complications. Mitral regurgitation was graded for severity by color Doppler flow imaging in all patients. Direction and size of systolic flow void in the left atrium were analyzed by contiguous multilevel cine MR images and the maximum volumes of flow void and left atrium were measured. Although flow void was found at the center of the left atrium in most of MSR, it was often directed along the postero-caudal atrial wall in anterior leaflet prolapse and along the anterocranial atrial wall in posterior leaflet prolapse. In MVP, the maximum volume of flow void was often seen in late systole. The maximum volume of flow void and that of left atrium were significantly larger in patients with atrial fibrillation and/or history of congestive heart failure. The length and volume of flow void were increased with clinical severity and degree of regurgitation determined by color Doppler flow imaging. Thus cine MR imaging provides a useful means for detection and semiquantitative evaluation of mitral regurgitation in subjects with MVP. (author).

  2. SU-E-T-335: Transit Dosimetry for Verification of Dose Delivery Using Electronic Portal Imaging Device (EPID)

    Energy Technology Data Exchange (ETDEWEB)

    Baek, T [Korea University, Seoul (Korea, Republic of); National Health Insurance Co.Ilsan Hospital, Ilsan (Korea, Republic of); Chung, E [National Health Insurance Co.Ilsan Hospital, Ilsan (Korea, Republic of); Lee, S [Cheil General Hospital and Women Healthcare Center, Kwandong University, Seoul (Korea, Republic of); Yoon, M [Korea University, Seoul (Korea, Republic of)

    2014-06-01

    Purpose: To evaluate the effectiveness of transit dose, measured with an electronic portal imaging device (EPID), in verifying actual dose delivery to patients. Methods: Plans of 5 patients with lung cancer, who received IMRT treatment, were examined using homogeneous solid water phantom and inhomogeneous anthropomorphic phantom. To simulate error in patient positioning, the anthropomorphic phantom was displaced from 5 mm to 10 mm in the inferior to superior (IS), superior to inferior (SI), left to right (LR), and right to left (RL) directions. The transit dose distribution was measured with EPID and was compared to the planed dose using gamma index. Results: Although the average passing rate based on gamma index (GI) with a 3% dose and a 3 mm distance-to-dose agreement tolerance limit was 94.34 % for the transit dose with homogeneous phantom, it was reduced to 84.63 % for the transit dose with inhomogeneous anthropomorphic phantom. The Result also shows that the setup error of 5mm (10mm) in IS, SI, LR and SI direction can Result in the decrease in values of GI passing rates by 1.3% (3.0%), 2.2% (4.3%), 5.9% (10.9%), and 8.9% (16.3%), respectively. Conclusion: Our feasibility study suggests that the transit dose-based quality assurance may provide information regarding accuracy of dose delivery as well as patient positioning.

  3. SPAMM, cine phase contrast imaging and fast spin-echo T2-weighted imaging in the study of intracranial cerebrospinal fluid (CSF) flow

    International Nuclear Information System (INIS)

    Connor, S.E.J.; O'Gorman, R.; Summers, P.; Simmons, A.; Moore, E.M.; Chandler, C.; Jarosz, J.M.

    2001-01-01

    AIM: To compare the qualitative assessment of cerebrospinal fluid (CSF) flow using a SPAMM (spatial modulation of magnetization) technique with cine phase contrast images (cine PC) and fast spin echo (FSE) T2-weighted images. MATERIALS AND METHODS: SPAMM, PC and T2-weighted sequences were performed on 22 occasions in 19 patients. Eleven of the studies were performed following a neuroendoscopic third ventriculostomy (NTV), and in these cases, the success of the NTV was determined by clinical follow-up. Two observers used consensus to grade the presence of CSF flow at nine different sites for each study. RESULTS: At 14 of the 178 matched sites, which could be assessed by both SPAMM and cine PC, SPAMM CSF flow grade was higher than that of cine PC. At a further 14/178 matched sites, the cine PC grade was higher than that of SPAMM. There was definite CSF flow at 113/182 (62%) of all the cine PC sites assessed, and 110/181 (61%) of all SPAMM sites assessed whilst 108/198 (54%) of FSE T2-weighted image sites demonstrated flow voids. Cine PC grades were higher than SPAMM at the cerebral aqueduct (P < 0.05, Wilcoxon sign rank test). Definite CSF flow within the anterior third ventricle was present in 4/5 (SPAMM) and 3/5 (cine PC) successful NTVs, 0/2 (SPAMM and cine PC) unsuccessful NTVs and 1/10 (SPAMM and cine PC) patients without NTV. CONCLUSION: SPAMM provides a comparable assessment of intracranial CSF flow to that of cine phase contrast imaging at all CSF sites except the cerebral aqueduct. Connor, S.E.J. et al. (2001)

  4. Right ventricular volume and mass determined by cine magnetic resonance imaging in HIV patients with possible right ventricular dysfunction

    DEFF Research Database (Denmark)

    Kjaer, Andreas; Lebech, Anne-Mette; Gerstoft, Jan

    2006-01-01

    Impaired right ventricular (RV) function has been reported to occur in patients with HIV when studied by echocardiography. However, for accurate evaluation of RV function and morphology, first-pass radionuclide ventriculography (RNV) and cine magnetic resonance imaging (cine-MRI) are methods...... ventricular ejection fraction (RVEF). To do so, we screened patients with RNV and performed an additional cine-MRI in those with reduced RVEF determined by RNV. Ninety patients with HIV were included. To evaluate the MRI measures exactly we included 18 age- and gender-matched healthy volunteers to establish...

  5. Investigations on uncertainties in patient positioning for prostate treatment with EPID; Untersuchungen zur Positionierungsgenauigkeit bei Prostatakonformationsbestrahlungen mittels Portal-Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Bakai, A.; Nuesslin, F. [Universitaetsklinik fuer Radioonkologie, Tuebingen (Germany). Abt. Medizinische Physik; Paulsen, F.; Plasswilm, L.; Bamberg, M. [Universitaetsklinik fuer Radioonkologie, Tuebingen (Germany). Abt. Strahlentherapie

    2002-02-01

    Background: Conformal radiotherapy techniques as used in prostate treatment allow to spare normal tissue by conforming the radiation fields to the shape of the planning target volume (PTV). To be able to fully utilize the advantages of these techniques correct patient positioning is an important prerequisite. This study employing an electronic portal imaging device (EPID) investigated the positioning uncertainties that occur in the pelvic region for different patient positioning devices. Patients and Methods: 15 patients with prostate cancer were irradiated with or without rectal balloon/pelvic mask at a linear accelerator with multileaf collimator (MLC). For each patient multiple portal images were taken from different directions and compared to the digitally reconstructed radiographs (DRRs) of the treatment planning system and to simulation films (Table 1, Figure 1). Results: In spite of different positioning devices, all patients showed comparable total positioning uncertainties of 4.0 mm (lateral), 4.5 mm (cranio-caudal) and 1.7 mm (dorso-ventral). The lateral positioning error was reduced for the pelvic mask patients while the cranio-caudal error increased (Table 2, Figure 2). A systematic and a random component sum up to the total positioning error, and a good estimate of the magnitudes of the two is possible from six to eight portal images (Figure 3). Conclusions: With a small number of portal images it is possible to find out the systematic and random positioning error of a patient. Knowledge of the random error can be used to resize the treatment margin which is clinically relevant since this error differs greatly for different patients (Figure 4). Image analysis with EPID is convenient, yet has some problems. For example, one only gets indirect information on the movement of the ventral rectum wall. The successful operation of positioning devices, although, needs further improvement - especially if one focuses on IMRT. (orig.) [German] Hintergrund

  6. Three-dimensional cardiac cine imaging using the kat ARC acceleration: Initial experience in clinical adult patients at 3T.

    Science.gov (United States)

    Okuda, Shigeo; Yamada, Yoshitake; Tanimoto, Akihiro; Fujita, Jun; Sano, Motoaki; Fukuda, Keiichi; Kuribayashi, Sachio; Jinzaki, Masahiro; Nozaki, Atsushi; Lai, Peng

    2015-09-01

    Three-dimensional cardiac cine imaging has demonstrated promising clinical 1.5-Tesla results; however, its application to 3T scanners has been limited because of the higher sensitivity to off-resonance artifacts. The aim of this study was to apply 3D cardiac cine imaging during a single breath hold in clinical patients on a 3T scanner using the kat ARC (k- and adaptive-t auto-calibrating reconstruction for Cartesian sampling) technique and to evaluate the interchangeability between 2D and 3D cine imaging for cardiac functional analysis and detection of abnormalities in regional wall motion. Following institutional review board approval, we obtained 2D cine images with an acceleration factor of two during multiple breath holds and 3D cine images with a net scan acceleration factor of 7.7 during a single breath hold in 20 patients using a 3T unit. Two readers independently evaluated the wall motion of the left ventricle (LV) using a 5-point scale, and the consistency in the detection of regional wall motion abnormality between 2D and 3D cine was analyzed by Cohen's kappa test. The LV volume was calculated at end-diastole and end-systole (LVEDV, LVESV); the ejection fraction (LVEF) and myocardial weight (LVmass) were also calculated. The relationship between functional parameters calculated for 2D and 3D cine images was analyzed using Pearson's correlation analysis. The bias and 95% limit of agreement (LA) were calculated using Bland-Altman plots. In addition, a qualitative evaluation of image quality was performed with regard to the myocardium-blood contrast, noise level and boundary definition. Despite slight degradation in image quality for 3D cine, excellent agreement was obtained for the detection of wall motion abnormalities between 2D and 3D cine images (κ=0.84 and 0.94 for each reader). Excellent correlations between the two imaging methods were shown for the evaluation of functional parameters (r>0.97). Slight differences in LVEDV, LVESV, LVEF and LVmass

  7. Accelerated three-dimensional cine phase contrast imaging using randomly undersampled echo planar imaging with compressed sensing reconstruction.

    Science.gov (United States)

    Basha, Tamer A; Akçakaya, Mehmet; Goddu, Beth; Berg, Sophie; Nezafat, Reza

    2015-01-01

    The aim of this study was to implement and evaluate an accelerated three-dimensional (3D) cine phase contrast MRI sequence by combining a randomly sampled 3D k-space acquisition sequence with an echo planar imaging (EPI) readout. An accelerated 3D cine phase contrast MRI sequence was implemented by combining EPI readout with randomly undersampled 3D k-space data suitable for compressed sensing (CS) reconstruction. The undersampled data were then reconstructed using low-dimensional structural self-learning and thresholding (LOST). 3D phase contrast MRI was acquired in 11 healthy adults using an overall acceleration of 7 (EPI factor of 3 and CS rate of 3). For comparison, a single two-dimensional (2D) cine phase contrast scan was also performed with sensitivity encoding (SENSE) rate 2 and approximately at the level of the pulmonary artery bifurcation. The stroke volume and mean velocity in both the ascending and descending aorta were measured and compared between two sequences using Bland-Altman plots. An average scan time of 3 min and 30 s, corresponding to an acceleration rate of 7, was achieved for 3D cine phase contrast scan with one direction flow encoding, voxel size of 2 × 2 × 3 mm(3) , foot-head coverage of 6 cm and temporal resolution of 30 ms. The mean velocity and stroke volume in both the ascending and descending aorta were statistically equivalent between the proposed 3D sequence and the standard 2D cine phase contrast sequence. The combination of EPI with a randomly undersampled 3D k-space sampling sequence using LOST reconstruction allows a seven-fold reduction in scan time of 3D cine phase contrast MRI without compromising blood flow quantification. Copyright © 2014 John Wiley & Sons, Ltd.

  8. Assessment of Left Ventricular Function and Mass on Free-Breathing Compressed Sensing Real-Time Cine Imaging.

    Science.gov (United States)

    Kido, Tomoyuki; Kido, Teruhito; Nakamura, Masashi; Watanabe, Kouki; Schmidt, Michaela; Forman, Christoph; Mochizuki, Teruhito

    2017-09-25

    Compressed sensing (CS) cine magnetic resonance imaging (MRI) has the advantage of being inherently insensitive to respiratory motion. This study compared the accuracy of free-breathing (FB) CS and breath-hold (BH) standard cine MRI for left ventricular (LV) volume assessment.Methods and Results:Sixty-three patients underwent cine MRI with both techniques. Both types of images were acquired in stacks of 8 short-axis slices (temporal/spatial resolution, 41 ms/1.7×1.7×6 mm 3 ) and compared for ejection fraction, end-diastolic and systolic volumes, stroke volume, and LV mass. Both BH standard and FB CS cine MRI provided acceptable image quality for LV volumetric analysis (score ≥3) in all patients (4.7±0.5 and 3.7±0.5, respectively; Pcine MRI (median, IQR: BH standard, 83.8 mL, 64.7-102.7 mL; FB CS, 79.0 mL, 66.0-101.0 mL; P=0.0006). The total acquisition times for BH standard and FB CS cine MRI were 113±7 s and 24±4 s, respectively (Pcine MRI is a clinically useful alternative to BH standard cine MRI in patients with impaired BH capacity.

  9. Cine magnetic resonance imaging, computed tomography and ultrasonography in the evaluation of chest wall invasion of lung cancer

    International Nuclear Information System (INIS)

    Yokozaki, Michiya; Nawano, Shigeru; Nagai, Kanji; Moriyama, Noriyuki; Kodama, Tetsuro; Nishiwaki, Yutaka.

    1997-01-01

    To assess the usefulness of cine-magnetic resonance imaging (cine-MRI) in the evaluation of chest wall invasion, we compared the results of cine-MRI with those of computed tomography (CT) and ultrasonography (US). Eleven patients were examined who had no pain and who were difficult to diagnose by routine examinations. MRI was performed with a Magnetom SP/4000, 1.5T unit (Siemens, Germany). For cine imaging, continuous turbo-FLUSH (ultra fast low angle shot) images were obtained at an orthogonal section to the chest wall during slow deep breathing. A CT scan was performed using a TCT 900S or Super Helix (Toshiba, Japan) at 1 cm intervals, with section thicknesses of 1 cm throughout the entire chest. US was performed with a model SSA-270A (Toshiba, Japan) with 7.5-MHz linear array scanners (PLF-705S; Toshiba, Japan). Sensitivity, specificity and accuracy were 67%, 75% and 73% for cine MRI, 67%, 63% and 64% for CT, 33%, 75% and 64% for US, respectively. These results indicate that cine MRI is potentially useful for the diagnosis of chest wall invasion of lung cancer. (author)

  10. Measurement of left ventricular volume by biplane cine magnetic resonance imaging in children

    Energy Technology Data Exchange (ETDEWEB)

    Ichida, Fukiko; Hamamichi, Yuuji; Hashimoto, Ikuo; Tsubata, Shinichi; Miyazaki, Ayumi; Okada, Toshio; Murakami, Arata; Futatsuya, Ryuusuke (Toyama Medical and Pharmaceutical Univ. (Japan))

    1993-09-01

    To determine the ability of cine magnetic resonance imaging (MRI) to assess left ventricular (LV) volumes, we studied 20 children (age 4 months to 10 years) with various heart disease, validated by comparison with biplane LV angiography. Previous MRI studies to assess LV volumes have used multiple axial planes, which are compromised by partial volume effects and are time consuming to acquire and analyze. Accordingly, an imaging approach using biplane cine MRI and planes aligned with the true cardiac axes (the intrinsic long and short axis) of the LV was developed in views comparable with biplane LV angiography. In all patients, LV volumes were calculated by a Simpson's rule algorithm, both in MRI and LV angiography. MRI determined LV volumes were slightly underestimated but correlated reasonably well with angiographic values (LVEDV: Y=0.88X + 1.58, R=0.98, LVESV: Y=0.72X + 1.02, R=0.98). Especially, even in the patients who have abnormal left ventricular geometry such as Tetralogy of Fallot, MRI determined LV volumes correlated well with angiographic values. It is concluded that biplane cine MRI, using the intrinsic LV long and short axis planes, permits noninvasive assessment of LV volumes in views comparable to standard angiographic projections and appears practical for clinical use in childhood heart disease, because the scan and analysis time are relatively short. (author).

  11. Normal nonuniformity of left ventricular contraction. Assessment by cine MR imaging with presaturation myocardial tagging

    International Nuclear Information System (INIS)

    Naito, H.; Arisawa, J.; Harada, K.; Yamagami, H.; Kozuka, T.; Tamura, S.

    1996-01-01

    Purpose: To identify the normal performance of left ventricular (LV) regional contraction using cine MR imaging with presaturation myocardial tagging. Material and Methods: Sixteen normal volunteers were examined on a 1.5 T MR system with tagging cine sequences. Tags were applied at end-diastole as 2 parallel black lines on short-axis and 4-chamber sections, and the fractional shortenings were calculated at 7 LV locations. Results: The following results were obtained with significance: A transmural gradient of contractility in the short-axis section; prolonged late-systolic endocardial shortening and epicardial early termination in the free wall; initial delay of shortening in the anterior wall; apical predominance of contractility; predominance of circumferential shortening in the free wall and of meridional shortening in the septum. These findings could be associated with myocardial fiber architecture, presumed wall stress and temporal asynergy of excitation. Conclusion: Cine MR imaging with myocardial tagging proved to be useful in assessing the nonuniformity of LV contraction. (orig.)

  12. Measurement of left ventricular volume by biplane cine magnetic resonance imaging in children

    Energy Technology Data Exchange (ETDEWEB)

    Ichida, Fukiko; Hamamichi, Yuuji; Hashimoto, Ikuo; Tsubata, Shinichi; Miyazaki, Ayumi; Okada, Toshio; Murakami, Arata; Futatsuya, Ryuusuke [Toyama Medical and Pharmaceutical Univ. (Japan)

    1993-09-01

    To determine the ability of cine magnetic resonance imaging (MRI) to assess left ventricular (LV) volumes, we studied 20 children (age 4 months to 10 years) with various heart disease, validated by comparison with biplane LV angiography. Previous MRI studies to assess LV volumes have used multiple axial planes, which are compromised by partial volume effects and are time consuming to acquire and analyze. Accordingly, an imaging approach using biplane cine MRI and planes aligned with the true cardiac axes (the intrinsic long and short axis) of the LV was developed in views comparable with biplane LV angiography. In all patients, LV volumes were calculated by a Simpson's rule algorithm, both in MRI and LV angiography. MRI determined LV volumes were slightly underestimated but correlated reasonably well with angiographic values (LVEDV: Y=0.88X + 1.58, R=0.98, LVESV: Y=0.72X + 1.02, R=0.98). Especially, even in the patients who have abnormal left ventricular geometry such as Tetralogy of Fallot, MRI determined LV volumes correlated well with angiographic values. It is concluded that biplane cine MRI, using the intrinsic LV long and short axis planes, permits noninvasive assessment of LV volumes in views comparable to standard angiographic projections and appears practical for clinical use in childhood heart disease, because the scan and analysis time are relatively short. (author).

  13. Measurement of left ventricular volume by biplane cine magnetic resonance imaging in children

    International Nuclear Information System (INIS)

    Ichida, Fukiko; Hamamichi, Yuuji; Hashimoto, Ikuo; Tsubata, Shinichi; Miyazaki, Ayumi; Okada, Toshio; Murakami, Arata; Futatsuya, Ryuusuke

    1993-01-01

    To determine the ability of cine magnetic resonance imaging (MRI) to assess left ventricular (LV) volumes, we studied 20 children (age 4 months to 10 years) with various heart disease, validated by comparison with biplane LV angiography. Previous MRI studies to assess LV volumes have used multiple axial planes, which are compromised by partial volume effects and are time consuming to acquire and analyze. Accordingly, an imaging approach using biplane cine MRI and planes aligned with the true cardiac axes (the intrinsic long and short axis) of the LV was developed in views comparable with biplane LV angiography. In all patients, LV volumes were calculated by a Simpson's rule algorithm, both in MRI and LV angiography. MRI determined LV volumes were slightly underestimated but correlated reasonably well with angiographic values (LVEDV: Y=0.88X + 1.58, R=0.98, LVESV: Y=0.72X + 1.02, R=0.98). Especially, even in the patients who have abnormal left ventricular geometry such as Tetralogy of Fallot, MRI determined LV volumes correlated well with angiographic values. It is concluded that biplane cine MRI, using the intrinsic LV long and short axis planes, permits noninvasive assessment of LV volumes in views comparable to standard angiographic projections and appears practical for clinical use in childhood heart disease, because the scan and analysis time are relatively short. (author)

  14. Cine MR imaging of valvular regurgitant flow: Correlations with a phantom study

    International Nuclear Information System (INIS)

    Evans, A.J.; Blinder, R.A.; Herfkens, R.J.; Spritzer, C.E.; Hedlund, L.W.; Podolak, M.J.

    1987-01-01

    Dynamic gradient echo MR images reveal that flow through diseased valves is usually associated with a decrease in signal intensity. The authors used a phantom to create turbulent flow through an orifice, analogous to a stenotic heart valve. Signal intensity decreased linearly ( r = .97) as a function of increase in Reynolds number. The area of signal loss also increased linearly ( r = .96) as the Reynolds number increased. The image appearance changed in an orderly fashion as turbulence and velocity increased. At low turbulence, flow is bright (velocity 123 cm/sec, Reynolds number >4,300). Similar findings have been observed in cine images of regurgitant lesions in cardiac patients

  15. Whole heart cine MR imaging of pulmonary veins in patients with congenital heart disease. Comparison with Spin Echo MR imaging

    International Nuclear Information System (INIS)

    Mitsui, Hideaki; Saito, Haruo; Ishibashi, Tadashi; Takahashi, Shoki; Zuguchi, Masayuki; Yamada, Shogo

    2002-01-01

    We evaluated the accuracy of Whole Heart Cine (WHC) magnetic resonance (MR) imaging in the depiction of pulmonary veins (PVs) in patients with congenital heart disease (CHD) compared to that of spin echo (SE) MR imaging. Among our 35 patients, 4 patients had anomalous PV return. Detectability of four PVs on each MR examination images were evaluated. MR imaging is an effective modality for the clarification of PVs, and WHC MR imaging is more useful in delineating PV anomalies than SE MR imaging. (author)

  16. Quantification of mechanical ventricular dyssynchrony. Direct comparison of velocity-encoded and cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Muellerleile, K.; Baholli, L.; Groth, M.

    2011-01-01

    Purpose: The preoperative assessment of mechanical dyssynchrony can help to improve patient selection in candidates for cardiac resynchronization therapy (CRT). The present study compared the performance of velocity-encoded (VENC) MRI to cine-magnetic resonance imaging (MRI) for quantifying mechanical ventricular dyssynchrony. Materials and Methods: VENC-MRI and cine-MRI were performed in 20 patients with heart failure NYHA class III and reduced ejection fraction (median: 24 %, interquartile range: 18 - 28 %) before CRT device implantation. The interventricular mechanical delay (IVMD) was assessed by VENC-MRI as the temporal difference between the onset of aortic and pulmonary flow. Intraventricular dyssynchrony was quantified by cine-MRI, using the standard deviation of time to maximal wall thickening in sixteen left ventricular segments (SDt-16). The response to CRT was assessed in a six-month follow-up. Results: 14 patients (70 %) clinically responded to CRT. A similar accuracy was found to predict the response to CRT by measurements of the IVMD and SDt-16 (75 vs. 70 %; p = ns). The time needed for data analysis was significantly shorter for the IVMD at 1.69 min (interquartile range: 1.66 - 1.88 min) compared to 9.63 min (interquartile range: 8.92 - 11.63 min) for the SDt-16 (p < 0.0001). Conclusion: Measurements of the IVMD by VENC-MRI and the SDt-16 by cine-MRI provide a similar accuracy to identify clinical responders to CRT. However, data analysis of the IVMD is significantly less time-consuming compared to data analysis of the SDt-16. (orig.)

  17. SU-F-T-486: A Simple Approach to Performing Light Versus Radiation Field Coincidence Quality Assurance Using An Electronic Portal Imaging Device (EPID)

    Energy Technology Data Exchange (ETDEWEB)

    Herchko, S; Ding, G [Vanderbilt University, Nashville, TN (United States)

    2016-06-15

    Purpose: To develop an accurate, straightforward, and user-independent method for performing light versus radiation field coincidence quality assurance utilizing EPID images, a simple phantom made of readily-accessible materials, and a free software program. Methods: A simple phantom consisting of a blocking tray, graph paper, and high-density wire was constructed. The phantom was used to accurately set the size of a desired light field and imaged on the electronic portal imaging device (EPID). A macro written for use in ImageJ, a free image processing software, was then use to determine the radiation field size utilizing the high density wires on the phantom for a pixel to distance calibration. The macro also performs an analysis on the measured radiation field utilizing the tolerances recommended in the AAPM Task Group #142. To verify the accuracy of this method, radiochromic film was used to qualitatively demonstrate agreement between the film and EPID results, and an additional ImageJ macro was used to quantitatively compare the radiation field sizes measured both with the EPID and film images. Results: The results of this technique were benchmarked against film measurements, which have been the gold standard for testing light versus radiation field coincidence. The agreement between this method and film measurements were within 0.5 mm. Conclusion: Due to the operator dependency associated with tracing light fields and measuring radiation fields by hand when using film, this method allows for a more accurate comparison between the light and radiation fields with minimal operator dependency. Removing the need for radiographic or radiochromic film also eliminates a reoccurring cost and increases procedural efficiency.

  18. SU-F-T-486: A Simple Approach to Performing Light Versus Radiation Field Coincidence Quality Assurance Using An Electronic Portal Imaging Device (EPID)

    International Nuclear Information System (INIS)

    Herchko, S; Ding, G

    2016-01-01

    Purpose: To develop an accurate, straightforward, and user-independent method for performing light versus radiation field coincidence quality assurance utilizing EPID images, a simple phantom made of readily-accessible materials, and a free software program. Methods: A simple phantom consisting of a blocking tray, graph paper, and high-density wire was constructed. The phantom was used to accurately set the size of a desired light field and imaged on the electronic portal imaging device (EPID). A macro written for use in ImageJ, a free image processing software, was then use to determine the radiation field size utilizing the high density wires on the phantom for a pixel to distance calibration. The macro also performs an analysis on the measured radiation field utilizing the tolerances recommended in the AAPM Task Group #142. To verify the accuracy of this method, radiochromic film was used to qualitatively demonstrate agreement between the film and EPID results, and an additional ImageJ macro was used to quantitatively compare the radiation field sizes measured both with the EPID and film images. Results: The results of this technique were benchmarked against film measurements, which have been the gold standard for testing light versus radiation field coincidence. The agreement between this method and film measurements were within 0.5 mm. Conclusion: Due to the operator dependency associated with tracing light fields and measuring radiation fields by hand when using film, this method allows for a more accurate comparison between the light and radiation fields with minimal operator dependency. Removing the need for radiographic or radiochromic film also eliminates a reoccurring cost and increases procedural efficiency.

  19. Development of automatic extraction method of left ventricular contours on long axis view MR cine images

    International Nuclear Information System (INIS)

    Utsunomiya, Shinichi; Iijima, Naoto; Yamasaki, Kazunari; Fujita, Akinori

    1995-01-01

    In the MRI cardiac function analysis, left ventricular volume curves and diagnosis parameters are obtained by extracting the left ventricular cavities as regions of interest (ROI) from long axis view MR cine images. The ROI extractions had to be done by manual operations, because automatization of the extraction is difficult. A long axis view left ventricular contour consists of a cardiac wall part and an aortic valve part. The above mentioned difficulty is due to the decline of contrast on the cardiac wall part, and the disappearance of edge on the aortic valve part. In this paper, we report a new automatic extraction method for long axis view MR cine images, which needs only 3 manually indicated points on the 1st image to extract all the contours from the total sequence of images. At first, candidate points of a contour are detected by edge detection. Then, selecting the best matched combination of candidate points by Dynamic Programming, the cardiac wall part is automatically extracted. The aortic valve part is manually extracted for the 1st image by indicating both the end points, and is automatically extracted for the rest of the images, by utilizing the aortic valve motion characteristics throughout a cardiac cycle. (author)

  20. Standardized cine-loop documentation in abdominal ultrasound facilitates offline image interpretation.

    Science.gov (United States)

    Dormagen, Johann Baptist; Gaarder, Mario; Drolsum, Anders

    2015-01-01

    One of the main disadvantages of conventional ultrasound is its operator dependency, which might impede the reproducibility of the sonographic findings. A new approach with cine-loops and standardized scan protocols can overcome this drawback. To compare abdominal ultrasound findings of immediate bedside reading by performing radiologist with offline reading by a non-performing radiologist, using standardized cine-loop sequences. Over a 6-month period, three radiologists performed 140 dynamic ultrasound organ-based examinations in 43 consecutive outpatients. Examination protocols were standardized and included predefined probe position and sequences of short cine-loops of the liver, gallbladder, pancreas, kidneys, and urine bladder, covering the organs completely in two planes. After bedside examinations, the studies were reviewed and read out immediately by the performing radiologist. Image quality was registered from 1 (no diagnostic value) to 5 (excellent cine-loop quality). Offline reading was performed blinded by a radiologist who had not performed the examination. Bedside and offline reading were compared with each other and with consensus results. In 140 examinations, consensus reading revealed 21 cases with renal disorders, 17 cases with liver and bile pathology, and four cases with bladder pathology. Overall inter-observer agreement was 0.73 (95% CI 0.61-0.91), with lowest agreement for findings of the urine bladder (0.36) and highest agreement in liver examinations (0.90). Disagreements between the two readings were seen in nine kidneys, three bladder examinations, one pancreas and bile system examinations each, and in one liver, giving a total number of mismatches of 11%. Nearly all cases of mismatch were of minor clinical significance. The median image quality was 3 (range, 2-5) with most examinations deemed a quality of 3. Compared to consensus reading, overall accuracy was 96% for bedside reading and 94% for offline reading. Standardized cine

  1. Fetal cardiac cine imaging using highly accelerated dynamic MRI with retrospective motion correction and outlier rejection.

    Science.gov (United States)

    van Amerom, Joshua F P; Lloyd, David F A; Price, Anthony N; Kuklisova Murgasova, Maria; Aljabar, Paul; Malik, Shaihan J; Lohezic, Maelene; Rutherford, Mary A; Pushparajah, Kuberan; Razavi, Reza; Hajnal, Joseph V

    2018-01-01

    Development of a MRI acquisition and reconstruction strategy to depict fetal cardiac anatomy in the presence of maternal and fetal motion. The proposed strategy involves i) acquisition and reconstruction of highly accelerated dynamic MRI, followed by image-based ii) cardiac synchronization, iii) motion correction, iv) outlier rejection, and finally v) cardiac cine reconstruction. Postprocessing entirely was automated, aside from a user-defined region of interest delineating the fetal heart. The method was evaluated in 30 mid- to late gestational age singleton pregnancies scanned without maternal breath-hold. The combination of complementary acquisition/reconstruction and correction/rejection steps in the pipeline served to improve the quality of the reconstructed 2D cine images, resulting in increased visibility of small, dynamic anatomical features. Artifact-free cine images successfully were produced in 36 of 39 acquired data sets; prolonged general fetal movements precluded processing of the remaining three data sets. The proposed method shows promise as a motion-tolerant framework to enable further detail in MRI studies of the fetal heart and great vessels. Processing data in image-space allowed for spatial and temporal operations to be applied to the fetal heart in isolation, separate from extraneous changes elsewhere in the field of view. Magn Reson Med 79:327-338, 2018. © 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. 2017 The Authors Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.

  2. Accuracy of accelerated cine MR imaging at 3 Tesla in longitudinal follow-up of cardiac function

    International Nuclear Information System (INIS)

    Sandner, Torleif A.; Huber, Armin M.; Theisen, Daniel; Reiser, Maximilian F.; Wintersperger, Bernd J.; Houck, Philip; Runge, Val M.; Sincleair, Spencer

    2008-01-01

    The ability of fast, parallel-imaging-based cine magnetic resonance (MR) to monitor global cardiac function in longitudinal exams at 3 Tesla was evaluated. Seventeen patients with chronic cardiac disease underwent serial cine MR imaging exams (n=3) at 3 Tesla. Data were acquired in short-axis orientation using cine steady-state free precession (SSFP) with a spatial resolution of 2.5 x 1.9 mm 2 at 45 ms temporal resolution. Multislice imaging (three slices/breath-hold) was performed using TSENSE acceleration (R=3) and standard single-slice cine (non-TSENSE) was performed at identical locations in consecutive breath-holds. End-diastolic volume (EDV), end-systolic volume (ESV), ejection fraction (EF) and myocardial mass (MM) of both cine approaches were compared for individual time-points as well as for longitudinal comparison. TSENSE-cine did not show significant differences for EDV (2.6 ml; P=.79), ESV (2.2 ml; P=0.81), EF (-0.3%; P=0.95) and MM (2.4 g; P=0.72) in comparison with non-TSENSE. Longitudinal ANOVA analysis did not reveal significant differences for any parameter, neither for non-TSENSE data (all P>0.7) nor for TSENSE data (all P>0.9). Multifactorial ANOVA showed non-significant differences (all P>0.7) at comparable data variances. Data acquisition was significantly shortened using TSENSE. Threefold accelerated multislice cine at 3 Tesla allows accurate assessment of volumetric LV data and accurate longitudinal monitoring of global LV function at a substantially shorter overall examination time. (orig.)

  3. Feasibility of megavoltage portal CT using an electronic portal imaging device (EPID) and a multi-level scheme algebraic reconstruction technique (MLS-ART)

    International Nuclear Information System (INIS)

    Guan, Huaiqun; Zhu, Yunping

    1998-01-01

    Although electronic portal imaging devices (EPIDs) are efficient tools for radiation therapy verification, they only provide images of overlapped anatomic structures. We investigated using a fluorescent screen/CCD-based EPID, coupled with a novel multi-level scheme algebraic reconstruction technique (MLS-ART), for a feasibility study of portal computed tomography (CT) reconstructions. The CT images might be useful for radiation treatment planning and verification. We used an EPID, set it to work at the linear dynamic range and collimated 6 MV photons from a linear accelerator to a slit beam of 1 cm wide and 25 cm long. We performed scans under a total of ∼200 monitor units (MUs) for several phantoms in which we varied the number of projections and MUs per projection. The reconstructed images demonstrated that using the new MLS-ART technique megavoltage portal CT with a total of 200 MUs can achieve a contrast detectibility of ∼2.5% (object size 5mmx5mm) and a spatial resolution of 2.5 mm. (author)

  4. Quantification of Esophageal Tumor Motion on Cine-Magnetic Resonance Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Lever, Frederiek M.; Lips, Irene M.; Crijns, Sjoerd P.M.; Reerink, Onne; Lier, Astrid L.H.M.W. van; Moerland, Marinus A.; Vulpen, Marco van; Meijer, Gert J., E-mail: g.j.meijer@umcutrecht.nl

    2014-02-01

    Purpose: To quantify the movement of esophageal tumors noninvasively on cine-magnetic resonance imaging (MRI) by use of a semiautomatic method to visualize tumor movement directly throughout multiple breathing cycles. Methods and Materials: Thirty-six patients with esophageal tumors underwent MRI. Tumors were located in the upper (8), middle (7), and lower (21) esophagus. Cine-MR images were collected in the coronal and sagittal plane during 60 seconds at a rate of 2 Hz. An adaptive correlation filter was used to automatically track a previously marked reference point. Tumor movement was measured in the craniocaudal (CC), left–right (LR), and anteroposterior (AP) directions and its relationship along the longitudinal axis of the esophagus was investigated. Results: Tumor registration within the individual images was typically done at a millisecond time scale. The mean (SD) peak-to-peak displacements in the CC, AP, and LR directions were 13.3 (5.2) mm, 4.9 (2.5) mm, and 2.7 (1.2) mm, respectively. The bandwidth to cover 95% of excursions from the mean position (c95) was also calculated to exclude outliers caused by sporadic movements. The mean (SD) c95 values were 10.1 (3.8) mm, 3.7 (1.9) mm, and 2.0 (0.9) mm in the CC, AP, and LR dimensions. The end-exhale phase provided a stable position in the respiratory cycle, compared with more variety in the end-inhale phase. Furthermore, lower tumors showed more movement than did higher tumors in the CC and AP directions. Conclusions: Intrafraction tumor movement was highly variable between patients. Tumor position proved the most stable during the respiratory cycle in the end-exhale phase. A better understanding of tumor motion makes it possible to individualize radiation delivery strategies accordingly. Cine-MRI is a successful noninvasive modality to analyze motion for this purpose in the future.

  5. Quantification of Esophageal Tumor Motion on Cine-Magnetic Resonance Imaging

    International Nuclear Information System (INIS)

    Lever, Frederiek M.; Lips, Irene M.; Crijns, Sjoerd P.M.; Reerink, Onne; Lier, Astrid L.H.M.W. van; Moerland, Marinus A.; Vulpen, Marco van; Meijer, Gert J.

    2014-01-01

    Purpose: To quantify the movement of esophageal tumors noninvasively on cine-magnetic resonance imaging (MRI) by use of a semiautomatic method to visualize tumor movement directly throughout multiple breathing cycles. Methods and Materials: Thirty-six patients with esophageal tumors underwent MRI. Tumors were located in the upper (8), middle (7), and lower (21) esophagus. Cine-MR images were collected in the coronal and sagittal plane during 60 seconds at a rate of 2 Hz. An adaptive correlation filter was used to automatically track a previously marked reference point. Tumor movement was measured in the craniocaudal (CC), left–right (LR), and anteroposterior (AP) directions and its relationship along the longitudinal axis of the esophagus was investigated. Results: Tumor registration within the individual images was typically done at a millisecond time scale. The mean (SD) peak-to-peak displacements in the CC, AP, and LR directions were 13.3 (5.2) mm, 4.9 (2.5) mm, and 2.7 (1.2) mm, respectively. The bandwidth to cover 95% of excursions from the mean position (c95) was also calculated to exclude outliers caused by sporadic movements. The mean (SD) c95 values were 10.1 (3.8) mm, 3.7 (1.9) mm, and 2.0 (0.9) mm in the CC, AP, and LR dimensions. The end-exhale phase provided a stable position in the respiratory cycle, compared with more variety in the end-inhale phase. Furthermore, lower tumors showed more movement than did higher tumors in the CC and AP directions. Conclusions: Intrafraction tumor movement was highly variable between patients. Tumor position proved the most stable during the respiratory cycle in the end-exhale phase. A better understanding of tumor motion makes it possible to individualize radiation delivery strategies accordingly. Cine-MRI is a successful noninvasive modality to analyze motion for this purpose in the future

  6. Fully automated segmentation of left ventricle using dual dynamic programming in cardiac cine MR images

    Science.gov (United States)

    Jiang, Luan; Ling, Shan; Li, Qiang

    2016-03-01

    Cardiovascular diseases are becoming a leading cause of death all over the world. The cardiac function could be evaluated by global and regional parameters of left ventricle (LV) of the heart. The purpose of this study is to develop and evaluate a fully automated scheme for segmentation of LV in short axis cardiac cine MR images. Our fully automated method consists of three major steps, i.e., LV localization, LV segmentation at end-diastolic phase, and LV segmentation propagation to the other phases. First, the maximum intensity projection image along the time phases of the midventricular slice, located at the center of the image, was calculated to locate the region of interest of LV. Based on the mean intensity of the roughly segmented blood pool in the midventricular slice at each phase, end-diastolic (ED) and end-systolic (ES) phases were determined. Second, the endocardial and epicardial boundaries of LV of each slice at ED phase were synchronously delineated by use of a dual dynamic programming technique. The external costs of the endocardial and epicardial boundaries were defined with the gradient values obtained from the original and enhanced images, respectively. Finally, with the advantages of the continuity of the boundaries of LV across adjacent phases, we propagated the LV segmentation from the ED phase to the other phases by use of dual dynamic programming technique. The preliminary results on 9 clinical cardiac cine MR cases show that the proposed method can obtain accurate segmentation of LV based on subjective evaluation.

  7. Cine Magnetic Resonance Imaging of the Small Bowel: Comparison of Different Oral Contrast Media

    International Nuclear Information System (INIS)

    Asbach, P.; Breitwieser, C.; Diederichs, G.; Eisele, S.; Kivelitz, D.; Taupitz, M.; Zeitz, M.; Hamm, B.; Klessen, C.

    2006-01-01

    Purpose: To evaluate several substances regarding small bowel distension and contrast on balanced steady-state free precession (bSSFP) cine magnetic resonance (MR) images. Material and Methods: Luminal contrast was evaluated in 24 volunteers after oral application of two different contrast agent groups leading to either bright lumen (pineapple, blueberry juice) or dark lumen (tap water, orange juice) on T1-weighted images. Bowel distension was evaluated in 30 patients ingesting either methylcellulose or mannitol solution for limiting intestinal absorption. Fifteen patients with duodeno-jejunal intubation served as the control. Quantitative evaluation included measurement of luminal signal intensities and diameters of four bowel segments, qualitative evaluation assessed luminal contrast and distension on a five-point scale. Results: Quantitative and qualitative evaluation of the four contrast agents revealed no significant differences regarding luminal contrast on bSSFP images. Quantitative evaluation revealed significantly lower (P<0.05) small bowel distension for three out of four segments (qualitative evaluation: two out of four segments) for methylcellulose in comparison to the control. Mannitol was found to be equal to the control. Conclusion: Oral ingestion of tap water or orange juice in combination with mannitol is recommended for cine MR imaging of the small bowel regarding luminal contrast and small bowel distension on bSSFP sequences

  8. Cine Magnetic Resonance Imaging of the Small Bowel: Comparison of Different Oral Contrast Media

    Energy Technology Data Exchange (ETDEWEB)

    Asbach, P.; Breitwieser, C.; Diederichs, G.; Eisele, S.; Kivelitz, D.; Taupitz, M.; Zeitz, M.; Hamm, B.; Klessen, C. [Charite - Universitatsmedizin Berlin, Charite Campus Mitte, Berlin (Germany). Dept. of Radiology

    2006-11-15

    Purpose: To evaluate several substances regarding small bowel distension and contrast on balanced steady-state free precession (bSSFP) cine magnetic resonance (MR) images. Material and Methods: Luminal contrast was evaluated in 24 volunteers after oral application of two different contrast agent groups leading to either bright lumen (pineapple, blueberry juice) or dark lumen (tap water, orange juice) on T1-weighted images. Bowel distension was evaluated in 30 patients ingesting either methylcellulose or mannitol solution for limiting intestinal absorption. Fifteen patients with duodeno-jejunal intubation served as the control. Quantitative evaluation included measurement of luminal signal intensities and diameters of four bowel segments, qualitative evaluation assessed luminal contrast and distension on a five-point scale. Results: Quantitative and qualitative evaluation of the four contrast agents revealed no significant differences regarding luminal contrast on bSSFP images. Quantitative evaluation revealed significantly lower (P<0.05) small bowel distension for three out of four segments (qualitative evaluation: two out of four segments) for methylcellulose in comparison to the control. Mannitol was found to be equal to the control. Conclusion: Oral ingestion of tap water or orange juice in combination with mannitol is recommended for cine MR imaging of the small bowel regarding luminal contrast and small bowel distension on bSSFP sequences.

  9. Real-time automatic fiducial marker tracking in low contrast cine-MV images

    Energy Technology Data Exchange (ETDEWEB)

    Lin, Wei-Yang; Lin, Shu-Fang; Yang, Sheng-Chang; Liou, Shu-Cheng; Nath, Ravinder; Liu Wu [Department of Computer Science and Information Engineering, National Chung Cheng University, Taiwan, 62102 (China); Department of Therapeutic Radiology, Yale University School of Medicine, New Haven, Connecticut 06510-3220 (United States)

    2013-01-15

    Purpose: To develop a real-time automatic method for tracking implanted radiographic markers in low-contrast cine-MV patient images used in image-guided radiation therapy (IGRT). Methods: Intrafraction motion tracking using radiotherapy beam-line MV images have gained some attention recently in IGRT because no additional imaging dose is introduced. However, MV images have much lower contrast than kV images, therefore a robust and automatic algorithm for marker detection in MV images is a prerequisite. Previous marker detection methods are all based on template matching or its derivatives. Template matching needs to match object shape that changes significantly for different implantation and projection angle. While these methods require a large number of templates to cover various situations, they are often forced to use a smaller number of templates to reduce the computation load because their methods all require exhaustive search in the region of interest. The authors solve this problem by synergetic use of modern but well-tested computer vision and artificial intelligence techniques; specifically the authors detect implanted markers utilizing discriminant analysis for initialization and use mean-shift feature space analysis for sequential tracking. This novel approach avoids exhaustive search by exploiting the temporal correlation between consecutive frames and makes it possible to perform more sophisticated detection at the beginning to improve the accuracy, followed by ultrafast sequential tracking after the initialization. The method was evaluated and validated using 1149 cine-MV images from two prostate IGRT patients and compared with manual marker detection results from six researchers. The average of the manual detection results is considered as the ground truth for comparisons. Results: The average root-mean-square errors of our real-time automatic tracking method from the ground truth are 1.9 and 2.1 pixels for the two patients (0.26 mm/pixel). The

  10. Real-time automatic fiducial marker tracking in low contrast cine-MV images

    International Nuclear Information System (INIS)

    Lin, Wei-Yang; Lin, Shu-Fang; Yang, Sheng-Chang; Liou, Shu-Cheng; Nath, Ravinder; Liu Wu

    2013-01-01

    Purpose: To develop a real-time automatic method for tracking implanted radiographic markers in low-contrast cine-MV patient images used in image-guided radiation therapy (IGRT). Methods: Intrafraction motion tracking using radiotherapy beam-line MV images have gained some attention recently in IGRT because no additional imaging dose is introduced. However, MV images have much lower contrast than kV images, therefore a robust and automatic algorithm for marker detection in MV images is a prerequisite. Previous marker detection methods are all based on template matching or its derivatives. Template matching needs to match object shape that changes significantly for different implantation and projection angle. While these methods require a large number of templates to cover various situations, they are often forced to use a smaller number of templates to reduce the computation load because their methods all require exhaustive search in the region of interest. The authors solve this problem by synergetic use of modern but well-tested computer vision and artificial intelligence techniques; specifically the authors detect implanted markers utilizing discriminant analysis for initialization and use mean-shift feature space analysis for sequential tracking. This novel approach avoids exhaustive search by exploiting the temporal correlation between consecutive frames and makes it possible to perform more sophisticated detection at the beginning to improve the accuracy, followed by ultrafast sequential tracking after the initialization. The method was evaluated and validated using 1149 cine-MV images from two prostate IGRT patients and compared with manual marker detection results from six researchers. The average of the manual detection results is considered as the ground truth for comparisons. Results: The average root-mean-square errors of our real-time automatic tracking method from the ground truth are 1.9 and 2.1 pixels for the two patients (0.26 mm/pixel). The

  11. Clinical advantages of three dimensional cine cardiac images

    International Nuclear Information System (INIS)

    Kinosada, Yasutomi; Okuda, Yasuyuki; Nakagawa, Tsuyoshi; Itou, Takafumi; Hattori, Takao.

    1996-01-01

    We evaluated clinical advantages and the quantitativeness of computerized three-dimensional (3D) cinematic images of a human heart, which were produced with a set of magnetic resonance (MR) images by using the computer graphic technique. Many contiguous, multi-location and multi-phase short axis images were obtained with the ECG gated conventional and fast cardiac imaging sequences in normal volunteers and selected patients with myocardial infarction, hypertrophic cardiomyopathy, dilated cardiomyopathy and left ventricular dysfunction. Judging by visual impressions of the computerized 3D cinematic cardiac images, we could easily understand and evaluate the myocardial motions or the anatomic and volumetric changes of a heart according to the cardiac phases. These images were especially useful to compare the wall motion, the left ventricular ejection-fraction (LVEF), or other cardiac functions and conditions between before and after therapeutic procedures such as percutaneous transluminal coronary angioplasty for patients with myocardial infarction. A good correlation between the LVEF calculated from a set of computerized 3D cinematic images and the ultra sound examinations were found. The results of our study showed that computerized 3D cinematic cardiac images were clinically useful to understand the myocardial motions qualitatively and to evaluate cardiac functions such as the LVEF quantitatively. (author)

  12. Application of cine cardiac MR imaging in normal subjects and patients with valvular, coronary artery, and aortic disease

    International Nuclear Information System (INIS)

    Maddahi, J.; Ostrzega, E.; Crues, J.; Honma, H.; Siegel, R.; Charuzi, Y.; Berman, D.

    1987-01-01

    Cine MR imaging was performed on 15 normal subjects and 27 patients with cardiac disease. In normal subjects, high signal intensity of flowing blood contrasted with that of the myocardium. In 16 patients with valvular regurgitation, signal void jet due to turbulence was visualized across the diseased valves. In three IHSS patients, thickened LV myocardium, mitral regurgitant jets, and systolic LV outflow jets were noted. Five patients with myocardial infarction (MI) showed thinning and/or hypokinesis of MI regions. In three patients with Marfan syndrome, aortic dilatation, insufficiency, and flap (one pt) were identified. Cine MR imaging is potentially useful for evaluation of a variety of cardiac diseases

  13. Texture analysis of cardiac cine magnetic resonance imaging to detect nonviable segments in patients with chronic myocardial infarction.

    Science.gov (United States)

    Larroza, Andrés; López-Lereu, María P; Monmeneu, José V; Gavara, Jose; Chorro, Francisco J; Bodí, Vicente; Moratal, David

    2018-04-01

    To investigate the ability of texture analysis to differentiate between infarcted nonviable, viable, and remote segments on cardiac cine magnetic resonance imaging (MRI). This retrospective study included 50 patients suffering chronic myocardial infarction. The data were randomly split into training (30 patients) and testing (20 patients) sets. The left ventricular myocardium was segmented according to the 17-segment model in both cine and late gadolinium enhancement (LGE) MRI. Infarcted myocardium regions were identified on LGE in short-axis views. Nonviable segments were identified as those showing LGE ≥ 50%, and viable segments those showing 0 cine images. A support vector machine (SVM) classifier was trained with different combination of texture features to obtain a model that provided optimal classification performance. The best classification on testing set was achieved with local binary patterns features using a 2D + t approach, in which the features are computed by including information of the time dimension available in cine sequences. The best overall area under the receiver operating characteristic curve (AUC) were: 0.849, sensitivity of 92% to detect nonviable segments, 72% to detect viable segments, and 85% to detect remote segments. Nonviable segments can be detected on cine MRI using texture analysis and this may be used as hypothesis for future research aiming to detect the infarcted myocardium by means of a gadolinium-free approach. © 2018 American Association of Physicists in Medicine.

  14. Left-ventricular reduction surgery: pre- and postoperative evaluation by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Kivelitz, D.E.; Enzweiler, C.N.H.; Wiese, T.H.; Lembcke, A.; Hamm, B.; Hotz, H.; Konertz, W.; Borges, A.C.; Baumann, G.

    2001-01-01

    Aim: To evaluate the role of cine magnetic resonance imaging (MRI) in the preoperative assessment and postoperative follow-up of patients undergoing left ventricular (LV) reduction surgery. Patients and Methods: 6 patients with cardiomegaly were examined on a 1.5 T MR imager before and after LV reduction surgery. The heart was imaged along the short and long axes using a breath-hold ECG-triggered cine gradient-echo sequence for assessing ventricular and valvular morphology and function and performing volumetry (end-diastolic and end-systolic volumes, ejection fraction). Results: Postoperatively, the mean ejection fraction increased from 21.7% to 33.4% and the enddiastolic and end-systolic left ventricular volumes decreased in all patients (304.0 and 252.5 ml before to 205.0 and 141.9 ml after surgery). Mean myocardial mass decreased slightly from 283.8 g to 242.7 g. Differences were significant for all parameters (p [de

  15. Visualization and quantitative analysis of the CSF pulsatile flow with cine MR phase imaging

    International Nuclear Information System (INIS)

    Katayama, Shinji; Itoh, Takahiko; Kinugasa, Kazushi; Asari, Shoji; Nishimoto, Akira; Tsuchida, Shohei; Ono, Atsushi; Ikezaki, Yoshikazu; Yoshitome, Eiji.

    1991-01-01

    The visualization and the quantitative analysis of the CSF pulsatile flow were performed on ten healthy volunteers with cine MR phase imaging, a combination of the phase-contrast technique and the cardiac-gating technique. The velocities appropriate for the visualization and the quantitative analysis of the CSF pulsatile flow were from 6.0 cm/sec to 15.0 cm/sec. The applicability of this method for the quantitative analysis was proven with a steady-flow phantom. Phase images clearly demonstrated a to-and-fro motion of the CSF flow in the anterior subarachnoid space and in the posterior subarachnoid space. The flow pattern of CSF on healthy volunteers depends on the cardiac cycle. In the anterior subarachnoid space, the cephalic CSF flow continued until a 70-msec delay after the R-wave of the ECG and then reversed to caudal. At 130-190 msec, the caudal CSF flow reached its maximum velocity; thereafter it reversed again to cephalic. The same turn appeared following the phase, but then the amplitude decreased. The cephalic peaked at 370-430 msec, while the caudal peaked at 490-550 msec. The flow pattern of the CSF flow in the posterior subarachnoid space was almost identical to that in the anterior subarachnoid space. Cine MR phase imaging is thus useful for the visualization and the quantitative analysis of the CSF pulsative flow. (author)

  16. Cine MRI of dissecting aneurysm

    International Nuclear Information System (INIS)

    Takaki, Hajime

    1991-01-01

    Cine MRI was performed in 25 cases of aortic dissection and comparative study among cine MRI, spin-echo static MRI, contrast-enhanced CT and intravenous digital subtraction angiography (IVDSA) was made. Cine MRI accurately detected aortic dissection. It was most accurate among various diagnostic methods in demonstration of entry site of dissection. Take-off of renal artery and its relation to true and false channels was also accurately demonstrated by cine MRI. The above results suggest that cine MRI can be an important diagnostic modality with almost equal diagnostic quality to those of conventional angiography. However, further technical improvement to shorten the imaging time seems necessary to replace angiography. (author)

  17. Automatic localization of the left ventricular blood pool centroid in short axis cardiac cine MR images.

    Science.gov (United States)

    Tan, Li Kuo; Liew, Yih Miin; Lim, Einly; Abdul Aziz, Yang Faridah; Chee, Kok Han; McLaughlin, Robert A

    2018-06-01

    In this paper, we develop and validate an open source, fully automatic algorithm to localize the left ventricular (LV) blood pool centroid in short axis cardiac cine MR images, enabling follow-on automated LV segmentation algorithms. The algorithm comprises four steps: (i) quantify motion to determine an initial region of interest surrounding the heart, (ii) identify potential 2D objects of interest using an intensity-based segmentation, (iii) assess contraction/expansion, circularity, and proximity to lung tissue to score all objects of interest in terms of their likelihood of constituting part of the LV, and (iv) aggregate the objects into connected groups and construct the final LV blood pool volume and centroid. This algorithm was tested against 1140 datasets from the Kaggle Second Annual Data Science Bowl, as well as 45 datasets from the STACOM 2009 Cardiac MR Left Ventricle Segmentation Challenge. Correct LV localization was confirmed in 97.3% of the datasets. The mean absolute error between the gold standard and localization centroids was 2.8 to 4.7 mm, or 12 to 22% of the average endocardial radius. Graphical abstract Fully automated localization of the left ventricular blood pool in short axis cardiac cine MR images.

  18. Functional cine MR imaging for the detection and mapping of intraabdominal adhesions: method and surgical correlation

    Energy Technology Data Exchange (ETDEWEB)

    Buhmann-Kirchhoff, Sonja; Reiser, Maximilian; Lienemann, Andreas [University Hospital Grosshadern, Ludwig-Maximilians-University Munich, Department of Clinical Radiology, Munich (Germany); Lang, Reinhold; Steitz, Heinrich O.; Jauch, Karl W. [University Hospital Munich-Grosshadern, Department of Surgery, Munich (Germany); Kirchhoff, Chlodwig [University Hospital Munich-Innenstadt, Department of Surgery, Munich (Germany)

    2008-06-15

    The purpose of this study was to evaluate the presence and localization of intraabdominal adhesions using functional cine magnetic resonance imaging (MRI) and to correlate the MR findings with intraoperative results. In a retrospective study, patients who had undergone previous abdominal surgery with suspected intraabdominal adhesions were examined. A true fast imaging with steady state precession sequence in transverse/sagittal orientation was used for a section-by-section dynamic depiction of visceral slide on a 1.5-Tesla system. After MRI, all patients underwent anew surgery. A nine-segment abdominal map was used to document the location and type of the adhesions. The intraoperative results were taken as standard of reference. Ninety patients were enrolled. During surgery 71 adhesions were detected, MRI depicted 68 intraabdominal adhesions. The most common type of adhesion in MRI was found between the anterior abdominal wall and small bowel loops (n = 22, 32.5%) and between small bowel loops and pelvic organs (n = 14, 20.6%). Comparing MRI with the intraoperative findings, sensitivity varied between 31 and 75% with a varying specificity between 65 and 92% in the different segments leading to an overall MRI accuracy of 89%. Functional cine MRI proved to be a useful examination technique for the identification of intraabdominal adhesions in patients with acute or chronic pain and corresponding clinical findings providing accurate results. However, no differentiation for symptomatic versus asymptomatic adhesions is possible. (orig.)

  19. CSF flow image using phase-contrast cine MR technique : preliminary clinical application

    International Nuclear Information System (INIS)

    Kim, Hyae Young; Choi, Hye Young; Baek, Seung Yeon; Lee, Sun Wha; Ko, Eun Joo; Lee, Myung Sook

    1997-01-01

    To evaluate the clinical usefulness of 2-D Cine PC (phase contrast) technique in visualizing the pattern and the site of abnormal CSF flow and to assess the effect of a third ventriculostomy in patients with hydrocephalus. The study group consisted of three normal controls and 13 patients with hydrocephalus, as shown on CT or MRI, and two patients who had undergone their third ventriculostomy. The technique was EKG-gated 2-D Cine PC MRI with velocity encoding 5cm/sec, TR 80msec, TE 12.3-15msec, and flip angle 15-60 degrees. Image quality was analyzed for variable sequences, and CSF flow was observed along the CSF flow pathway. We analyzed continuity and intensity of the CSF flow signal, and obstruction site and flow velocity degree were then defined. Systolic high and diastolic low signal intensity along the CSF flow-pathway, with normal asynchronicity and continuation, were clearly seen in normal controls. In three patients, there was obstruction at the ventricular level while others were either normal or showed a normal pattern with a weak signal. 'Normal' was defined as noncommunicating hydrocephalus and the latter as communicating hydrocephalus. In the two patients who had undergone ventriculostomy, a signal was in one case detected at the site of the third operation. A 2-D Cine PC CSF flow study enables us to see CSF flow signals noninvasively and to detect the site of obstruction of a CSF flow-pathway. It can therefore it can be useful for determining the application of a ventriculoperitoneal shunt and assessing the effect of a third ventriculostomy

  20. Area detection of uterine peristalsis using cine-MR images

    International Nuclear Information System (INIS)

    Sato, Tetsuo; Fujita, Nao; Nakai, Asako; Togashi, Kaori; Kuhara, Shigehide; Okada, Tomohisa

    2011-01-01

    In this paper, a technique for evaluating uterine peristalsis obtained by magnetic resonance images is proposed. Uterine peristalsis is the wavelike movement of uterine muscle contractions with rhythm and direction. Correlation between direction of uterine peristalsis and menstrual cycle has been reported, which is supposed to help sperm transpotation and implantation. While evaluation of uterine peristalsis has been done, they are very subjective and no quantitative evaluation has been established. In the proposed method the peristalsis area and their contraction intensity were calculated. The points of uterine peristalsis was estimated by the spatio-temporal map, which reflect the time series behavior of the intensities around points set on uterine endometrial boundaries. Then peristalsis area was decided as the estimated points. The contraction intensity was defined as the ratio of the signal intensity change in the peristalsis area. (author)

  1. Comparative assessment of liver tumor motion using cine-magnetic resonance imaging versus 4-dimensional computed tomography.

    Science.gov (United States)

    Fernandes, Annemarie T; Apisarnthanarax, Smith; Yin, Lingshu; Zou, Wei; Rosen, Mark; Plastaras, John P; Ben-Josef, Edgar; Metz, James M; Teo, Boon-Keng

    2015-04-01

    To compare the extent of tumor motion between 4-dimensional CT (4DCT) and cine-MRI in patients with hepatic tumors treated with radiation therapy. Patients with liver tumors who underwent 4DCT and 2-dimensional biplanar cine-MRI scans during simulation were retrospectively reviewed to determine the extent of target motion in the superior-inferior, anterior-posterior, and lateral directions. Cine-MRI was performed over 5 minutes. Tumor motion from MRI was determined by tracking the centroid of the gross tumor volume using deformable image registration. Motion estimates from 4DCT were performed by evaluation of the fiducial, residual contrast (or liver contour) positions in each CT phase. Sixteen patients with hepatocellular carcinoma (n=11), cholangiocarcinoma (n=3), and liver metastasis (n=2) were reviewed. Cine-MRI motion was larger than 4DCT for the superior-inferior direction in 50% of patients by a median of 3.0 mm (range, 1.5-7 mm), the anterior-posterior direction in 44% of patients by a median of 2.5 mm (range, 1-5.5 mm), and laterally in 63% of patients by a median of 1.1 mm (range, 0.2-4.5 mm). Cine-MRI frequently detects larger differences in hepatic intrafraction tumor motion when compared with 4DCT most notably in the superior-inferior direction, and may be useful when assessing the need for or treating without respiratory management, particularly in patients with unreliable 4DCT imaging. Margins wider than the internal target volume as defined by 4DCT were required to encompass nearly all the motion detected by cine-MRI for some of the patients in this study. Copyright © 2015 Elsevier Inc. All rights reserved.

  2. A four-dimensional motion field atlas of the tongue from tagged and cine magnetic resonance imaging

    Science.gov (United States)

    Xing, Fangxu; Prince, Jerry L.; Stone, Maureen; Wedeen, Van J.; El Fakhri, Georges; Woo, Jonghye

    2017-02-01

    Representation of human tongue motion using three-dimensional vector fields over time can be used to better understand tongue function during speech, swallowing, and other lingual behaviors. To characterize the inter-subject variability of the tongue's shape and motion of a population carrying out one of these functions it is desirable to build a statistical model of the four-dimensional (4D) tongue. In this paper, we propose a method to construct a spatio-temporal atlas of tongue motion using magnetic resonance (MR) images acquired from fourteen healthy human subjects. First, cine MR images revealing the anatomical features of the tongue are used to construct a 4D intensity image atlas. Second, tagged MR images acquired to capture internal motion are used to compute a dense motion field at each time frame using a phase-based motion tracking method. Third, motion fields from each subject are pulled back to the cine atlas space using the deformation fields computed during the cine atlas construction. Finally, a spatio-temporal motion field atlas is created to show a sequence of mean motion fields and their inter-subject variation. The quality of the atlas was evaluated by deforming cine images in the atlas space. Comparison between deformed and original cine images showed high correspondence. The proposed method provides a quantitative representation to observe the commonality and variability of the tongue motion field for the first time, and shows potential in evaluation of common properties such as strains and other tensors based on motion fields.

  3. Evaluation of cardiac motion and function by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Kondo, Takeshi; Kurokawa, Hiroshi; Anno, Hirofumi

    1992-01-01

    Cardiac cine magnetic resonance imaging (MRI) was studied to evaluate the cardiac motion and function, and a water-stream phantom study was performed to clarify whether it was possible to quantitatively assess the valvular regurgitation flow by the size of the flow void. In normal subjects, the left ventricular (LV) epicardial apex swung up to the base only a few millimeters, and the mitral annulus ring moved about 14 mm as mean value toward the apex during systole. Those motions of mitral annulus ring may contribute to the left atrial filling. The LV longitudinal shortening and torsions were shown by the tagging method. This tagging method was the best method for estimating cardiac motions. Cardiac cine MRI using software including a modified Simpson's method program and a wall motion analysis program was useful for routine LV volumetry and wall motion analysis because it was a simple and reliable method. Our water-stream phantom studies demonstrated that it might be difficult to perform quantitative evaluation of valvular regurgitation flow by using only the size of the flow void without acquiring information relating to the orifice area. (author)

  4. MR imaging of the pulmonary vasculature: Cine and high-resolution techniques

    International Nuclear Information System (INIS)

    Gefter, W.B.; Hatabu, H.; Kressel, H.Y.; Axel, L.; Lenkinski, R.E.; Schiebler, M.L.; Dougherty, L.; Douglas, P.S.; Reichek, N.

    1987-01-01

    Pulmonary vessels were evaluated on 43 cine examinations (12 normals, 31 with cardiopulmonary diseases) at 1.5 T (General Electric). Arteries and veins could be differentiated by characteristic intensity fluctuations in 90%. Abnormal patterns were observed with elevated left atrial pressure, pulmonary hypertension, pulmonic stenosis, and mitral regurgitation. A small arteriovenous malformation was identified. Approaches to high-resolution imaging included surface coils, 24-cm field of view, and 256 x 256 matrix. Spin-echo (SE) sequences gated in systole or diastole, and GRASS with and without breath-holding were evaluated. Surface-coil SE diastolic images (4 NEX) visualized sixth- and seventh-generation vessels. Breath-hold GRASS showed fifth- and sixth-generation vessels without respiratory artifact. These are promising techniques for displaying the pulmonary circulation

  5. Uterine contractions evaluated on cine MR imaging in patients with uterine leiomyomas

    International Nuclear Information System (INIS)

    Nishino, Mizuki; Togashi, Kaori; Nakai, Asako; Hayakawa, Katsumi; Kanao, Shotarou; Iwasaku, Kazuhiro; Fujii, Shingo

    2005-01-01

    Purpose: Submucosal leiomyoma is one of the most recognized causes of infertility and habitual abortion. The purpose of this study is to evaluate uterine peristalsis, a cycle-related inherent contractility of uterus probably responsible for sperm transport and conservation of pregnancy, in patients with uterine leiomyomas using cine magnetic resonance (MR) imaging. Materials and methods: Study population consisted of 26 female patients (age range: 19-51 years, mean: 41 years), in which 16 patients had submucosal leiomyomas and 10 patients had intramural or subserosal leiomyomas. We prospectively performed MR imaging of the midsagittal plane of uterus using 1.5 T magnet (Symphony, Siemens Medical Systems) with a body array coil, and obtained 60 half-Fourier acquisition single shot turbo spin echo (HASTE) images (Echo time=80 ms, FOV=300 mm, slice thickness 5 mm, matrix 256x256) within 2 min, and displayed them on cine mode at 12x faster than real speed. Evaluated were peristaltic movements at the endometral-myometrial junction and focal myometrial movements, adjacent to leiomyomas, regarding presence, direction, frequency, and conduction. Results: The peristaltic movements were identified in 12/16 patients with submucosal lesions and 10/10 with other leiomyomas. The frequency and direction were cycle-related. Loss of peristalsis was noted adjacent to submucosal myomas in 4/12 patients, but was not in others. Focal myometrial movements were noted in 9/16 patients with submucosal myomas, but not in others. Conclusions: Uterine peristaltic movements were partly interrupted by submucosal leiomoymas, but not by myometrial or subserosal leiomyomas. Loss of peristalsis and focal myometrial movements was noted only adjacent to submucosal leiomyomas. These findings are considered to represent dysfunctional contractility, and may be related with pregnancy loss

  6. Uterine contractions evaluated on cine MR imaging in patients with uterine leiomyomas

    Energy Technology Data Exchange (ETDEWEB)

    Nishino, Mizuki E-mail: mizuki@mbox.kyoto-inet.or.jpnishinomizuki@hotmail.com; Togashi, Kaori; Nakai, Asako; Hayakawa, Katsumi; Kanao, Shotarou; Iwasaku, Kazuhiro; Fujii, Shingo

    2005-01-01

    Purpose: Submucosal leiomyoma is one of the most recognized causes of infertility and habitual abortion. The purpose of this study is to evaluate uterine peristalsis, a cycle-related inherent contractility of uterus probably responsible for sperm transport and conservation of pregnancy, in patients with uterine leiomyomas using cine magnetic resonance (MR) imaging. Materials and methods: Study population consisted of 26 female patients (age range: 19-51 years, mean: 41 years), in which 16 patients had submucosal leiomyomas and 10 patients had intramural or subserosal leiomyomas. We prospectively performed MR imaging of the midsagittal plane of uterus using 1.5 T magnet (Symphony, Siemens Medical Systems) with a body array coil, and obtained 60 half-Fourier acquisition single shot turbo spin echo (HASTE) images (Echo time=80 ms, FOV=300 mm, slice thickness 5 mm, matrix 256x256) within 2 min, and displayed them on cine mode at 12x faster than real speed. Evaluated were peristaltic movements at the endometral-myometrial junction and focal myometrial movements, adjacent to leiomyomas, regarding presence, direction, frequency, and conduction. Results: The peristaltic movements were identified in 12/16 patients with submucosal lesions and 10/10 with other leiomyomas. The frequency and direction were cycle-related. Loss of peristalsis was noted adjacent to submucosal myomas in 4/12 patients, but was not in others. Focal myometrial movements were noted in 9/16 patients with submucosal myomas, but not in others. Conclusions: Uterine peristaltic movements were partly interrupted by submucosal leiomoymas, but not by myometrial or subserosal leiomyomas. Loss of peristalsis and focal myometrial movements was noted only adjacent to submucosal leiomyomas. These findings are considered to represent dysfunctional contractility, and may be related with pregnancy loss.

  7. A trial to reduce cardiac motion artifact on HR-CT images of the lung with the use of subsecond scan and special cine reconstruction algorithm

    Energy Technology Data Exchange (ETDEWEB)

    Sakai, Fumikazu; Tsuuchi, Yasuhiko; Suzuki, Keiko; Ueno, Keiko; Yamada, Takayuki; Okawa, Tomohiko [Tokyo Women`s Medical Coll. (Japan); Yun, Shen; Horiuchi, Tetsuya; Kimura, Fumiko

    1998-05-01

    We describe our trial to reduce cardiac motion artifacts on HR-CT images caused by cardiac pulsation by combining use of subsecond CT (scan time 0.8 s) and a special cine reconstruction algorithm (cine reconstruction algorithm with 180-degree helical interpolation). Eleven to 51 HR-CT images were reconstructed with the special cine reconstruction algorithm at the pitch of 0.1 (0.08 s) from the data obtained by two to six contigious rotation scans at the same level. Images with the fewest cardiac motion artifacts were selected for evaluation. These images were compared with those reconstructed with a conventional cine reconstruction algorithm and step-by-step scan. In spite of its increased radiation exposure, technical complexity and slight degradation of spatial resolution, our method was useful in reducing cardiac motion artifacts on HR-CT images in regions adjacent to the heart. (author)

  8. Spirometer-controlled cine magnetic resonance imaging to diagnose tracheobronchomalacia in pediatric patients

    DEFF Research Database (Denmark)

    Ciet, Pierluigi; Wielopolski, Piotr; Manniesing, Rashindra

    2014-01-01

    is restricted by ionizing radiation. Our aim was to evaluate the feasibility of spirometer-controlled cine-MRI as alternative to cine-CT in a retrospective study.12 children (mean 12 years, range 7-17), suspected to have TBM, underwent cine-MRI. Static scans were acquired at end-inspiration and expiration......Tracheobronchomalacia (TBM) is defined as an excessive collapse of the intrathoracic trachea. Bronchoscopy is the gold standard to diagnose TBM, but bronchoscopy has major disadvantages, such as general anaesthesia. Cine-CT is a non-invasive alternative to diagnose TBM, but its use in children...

  9. Cine MR imaging assessment of regional left ventricular systolic wall thickening in patients with remote myocardial infarction

    International Nuclear Information System (INIS)

    Pfugfelder, P.; White, R.D.; Sechtem, U.; Gould, R.G.; Higgins, C.B.

    1986-01-01

    Cine MR imaging, a new rapid imaging technique, was used to acquire transverse images of the heart at a rate of 16-30 frames per cardiac cycle. Left ventricular wall thickness was measured at end diastole and end systole in six regions in the midventricular section of 13 healthy subjects and seven patients with previously documented myocardial infarction. Mean percent systolic wall thickening (%SWT) was 51% +- 26% in healthy subjects. In patients, %SWT was -8% +- 22% in the infarct zone and 42% +- 22% in the normal myocardium. In addition to the qualitative information derived from the cinematic display, determination of regional %SWT by cine-MR imaging may be useful for quantifying regional left ventricular dysfunction

  10. Quantitative measurement of normal and hydrocephalic cerebrospinal fluid flow using phase contrast cine MR imaging

    International Nuclear Information System (INIS)

    Katayama, Shinji; Asari, Shoji; Ohmoto, Takashi

    1993-01-01

    Measurements of the cerebrospinal fluid (CSF) flow using phase contrast cine magnetic resonance (MR) imaging were performed on a phantom, 12 normal subjects and 20 patients with normal pressure hydrocephalus (NPH). The phantom study demonstrated the applicability of phase contrast in quantitative measurement of the slow flow. The CSF flows of the normal subjects showed a consistent pattern with a to-and-fro movement of the flow in the anterior subarachnoid space at the C2/3 level, and they were dependent on the cardiac cycle in all subjects. However, the patients with NPH showed variable patterns of the CSF pulsatile flow and these patterns could be divided into four types according to velocity and amplitude. The amplitudes of each type were as follows: type 0 (n=1), 87.6 mm; type I (n=2), 58.2 mm (mean); type II (n=6), 48.0±5.0 mm (mean±SEM); and type III (n=11), 19.9±1.8 mm (mean±SEM). The decrease of the amplitudes correlated to a worsening of the clinical symptoms. After the shunting operation, the amplitude of to-and-fro movement of the CSF increased again in the patients with NPH who improved clinically. Some of the type III cases were reclassified type II, I and 0 and also one of the type II cases changed type I after the shunting operation. We conclude that the phase contrast cine MR imaging is a practically and clinically applicable technique for the quantitative measurement of the CSF flow. (author)

  11. Evaluation of ventricular dysfunction using semi-automatic longitudinal strain analysis of four-chamber cine MR imaging.

    Science.gov (United States)

    Kawakubo, Masateru; Nagao, Michinobu; Kumazawa, Seiji; Yamasaki, Yuzo; Chishaki, Akiko S; Nakamura, Yasuhiko; Honda, Hiroshi; Morishita, Junji

    2016-02-01

    The aim of this study was to evaluate ventricular dysfunction using the longitudinal strain analysis in 4-chamber (4CH) cine MR imaging, and to investigate the agreement between the semi-automatic and manual measurements in the analysis. Fifty-two consecutive patients with ischemic, or non-ischemic cardiomyopathy and repaired tetralogy of Fallot who underwent cardiac MR examination incorporating cine MR imaging were retrospectively enrolled. The LV and RV longitudinal strain values were obtained by semi-automatically and manually. Receiver operating characteristic (ROC) analysis was performed to determine the optimal cutoff of the minimum longitudinal strain value for the detection of patients with cardiac dysfunction. The correlations between manual and semi-automatic measurements for LV and RV walls were analyzed by Pearson coefficient analysis. ROC analysis demonstrated the optimal cut-off of the minimum longitudinal strain values (εL_min) for diagnoses the LV and RV dysfunction at a high accuracy (LV εL_min = -7.8 %: area under the curve, 0.89; sensitivity, 83 %; specificity, 91 %, RV εL_min = -15.7 %: area under the curve, 0.82; sensitivity, 92 %; specificity, 68 %). Excellent correlations between manual and semi-automatic measurements for LV and RV free wall were observed (LV, r = 0.97, p cine MR imaging can evaluate LV and RV dysfunction with simply and easy measurements. The strain analysis could have extensive application in cardiac imaging for various clinical cases.

  12. Cine-MR imaging in determining the flow characteristics of CSF and blood in spinal and intracranial lesions

    International Nuclear Information System (INIS)

    Post, M.J.D.; Quencer, R.M.; Green, B.A.; Hinks, R.S.; Sklar, E.M.L.; Patchen, S.J.

    1988-01-01

    The purpose of this prospective study was to determine the value of cine magnetic resonance (MR) imaging in assessing the flow patterns of patients with vascular cord neoplasms, spinal cord and subarachnoid cysts, obstructive hydrocephalus, and intracranial aneurysms. The authors' results in 26 patients showed that cine MR imaging can be used to (1) identify spinal neoplasms with prominent vascular supply; (2) help distinguish spinal cord cysts occurring above a spinal cord tumor from tumoral cysts; (3) determine which spinal cord or subarachnois cysts need shunting and, postoperatively, which cysts are adequately decompressed; (4) establish which tonsillar herniations in Chiari malformations may require resection; (5) determine the site of the block in obstructive hydrocephalus; and (6) determine the flow characteristics of an aneurysm

  13. Left ventricular volume measurements with free breathing respiratory self-gated 3-dimensional golden angle radial whole-heart cine imaging - Feasibility and reproducibility.

    Science.gov (United States)

    Holst, Karen; Ugander, Martin; Sigfridsson, Andreas

    2017-11-01

    To develop and evaluate a free breathing respiratory self-gated isotropic resolution technique for left ventricular (LV) volume measurements. A 3D radial trajectory with double golden-angle ordering was used for free-running data acquisition during free breathing in 9 healthy volunteers. A respiratory self-gating signal was extracted from the center of k-space and used with the electrocardiogram to bin all data into 3 respiratory and 25 cardiac phases. 3D image volumes were reconstructed and the LV endocardial border was segmented. LV volume measurements and reproducibility from 3D free breathing cine were compared to conventional 2D breath-held cine. No difference was found between 3D free breathing cine and 2D breath-held cine with regards to LV ejection fraction, stroke volume, end-systolic volume and end-diastolic volume (Pcine and 2D breath-held cine (Pcine and conventional 2D breath-held cine showed similar values and test-retest repeatability for LV volumes in healthy volunteers. 3D free breathing cine enabled retrospective sorting and arbitrary angulation of isotropic data, and could correctly measure LV volumes during free breathing acquisition. Copyright © 2017 Elsevier Inc. All rights reserved.

  14. Retrospective Reconstruction of High Temporal Resolution Cine Images from Real-Time MRI using Iterative Motion Correction

    DEFF Research Database (Denmark)

    Hansen, Michael Schacht; Sørensen, Thomas Sangild; Arai, Andrew

    2012-01-01

    acquisitions in 10 (N = 10) subjects. Acceptable image quality was obtained in all motion-corrected reconstructions, and the resulting mean image quality score was (a) Cartesian real-time: 2.48, (b) Golden Angle real-time: 1.90 (1.00–2.50), (c) Cartesian motion correction: 3.92, (d) Radial motion correction: 4...... and motion correction based on nonrigid registration and can be applied to arbitrary k-space trajectories. The method is demonstrated with real-time Cartesian imaging and Golden Angle radial acquisitions, and the motion-corrected acquisitions are compared with raw real-time images and breath-hold cine...

  15. Cine magnetic resonance

    International Nuclear Information System (INIS)

    Higgins, C.B.; Sechtem, U.P.; Pflugfelder, P.

    1987-01-01

    Cine magnetic resonance (MR) is a fast MR imaging process with referencing of the imaging data to the electrocardiogram (ECG) so that images corresponding to 21-msec segments of the cardiac cycle are acquired. A series of such images, each corresponding to a 21-msec segment of the cardiac cycle, can be laced together for viewing in the cine format at a framing rate of 20 to 40 frames per second. Since cine angiograms of the heart are usually done at 30 frames per second, this technique achieves a temporal resolution adequate for the evluation of central cardiovascular function. The major application of this technique is to depict central cardiovascular function and blood flow

  16. Feasibility study on 3D image reconstruction from 2D orthogonal cine-MRI for MRI-guided radiotherapy.

    Science.gov (United States)

    Paganelli, Chiara; Lee, Danny; Kipritidis, John; Whelan, Brendan; Greer, Peter B; Baroni, Guido; Riboldi, Marco; Keall, Paul

    2018-02-11

    In-room MRI is a promising image guidance strategy in external beam radiotherapy to acquire volumetric information for moving targets. However, limitations in spatio-temporal resolution led several authors to use 2D orthogonal images for guidance. The aim of this work is to present a method to concurrently compensate for non-rigid tumour motion and provide an approach for 3D reconstruction from 2D orthogonal cine-MRI slices for MRI-guided treatments. Free-breathing sagittal/coronal interleaved 2D cine-MRI were acquired in addition to a pre-treatment 3D volume in two patients. We performed deformable image registration (DIR) between cine-MRI slices and corresponding slices in the pre-treatment 3D volume. Based on an extrapolation of the interleaved 2D motion fields, the 3D motion field was estimated and used to warp the pre-treatment volume. Due to the lack of a ground truth for patients, the method was validated on a digital 4D lung phantom. On the phantom, the 3D reconstruction method was able to compensate for tumour motion and compared favourably to the results of previously adopted strategies. The difference in the 3D motion fields between the phantom and the extrapolated motion was 0.4 ± 0.3 mm for tumour and 0.8 ± 1.5 mm for whole anatomy, demonstrating feasibility of performing a 3D volumetric reconstruction directly from 2D orthogonal cine-MRI slices. Application of the method to patient data confirmed the feasibility of utilizing this method in real world scenarios. Preliminary results on phantom and patient cases confirm the feasibility of the proposed approach in an MRI-guided scenario, especially for non-rigid tumour motion compensation. © 2018 The Royal Australian and New Zealand College of Radiologists.

  17. Assessment of left ventricular mass in sequential studies with cine MR imaging

    International Nuclear Information System (INIS)

    Tomei, E.; Semelka, R.; Wagner, S.; Mayo, J.; Chatterjee, K.; Parmley, W.W.; O'Sullivan, M.; Wolfe, C.L.; Caputo, G.; Higgins, C.B.

    1989-01-01

    The aim of this study was to measure left ventricular (LV) mass in 11 healthy volunteers, 10 patients with dilated cardiomyopathy, and eight patients with LV hypertrophy (LVH), using two sequential studies to compare the characteristics of LV mass in the same subject and in different clinical situations. All subjects underwent short- axis cine MR imaging at 1.5 T. Each subject had two separate studies: the healthy volunteers within 6 months and those with cardiac disease within 1 week. Measurements (both end- systolic and end-diastolic) included LV mass, LV mass index, and wall thickness. LV mass was substantially increased in both DCM and LVH. The interstudy variability for end- systolic and end-diastolic mass was 5.2% and 3.8%, respectively, for healthy volunteers, 5.2% and 4.0% for LVH, and 3.8% and 6.1% for DCM. The low variability indicates the reproducibility of this technique in sequential studies when no change is expected

  18. SU-G-BRA-03: PCA Based Imaging Angle Optimization for 2D Cine MRI Based Radiotherapy Guidance

    Energy Technology Data Exchange (ETDEWEB)

    Chen, T; Yue, N; Jabbour, S; Zhang, M [Rutgers University, New Brunswick, NJ (United States)

    2016-06-15

    Purpose: To develop an imaging angle optimization methodology for orthogonal 2D cine MRI based radiotherapy guidance using Principal Component Analysis (PCA) of target motion retrieved from 4DCT. Methods: We retrospectively analyzed 4DCT of 6 patients with lung tumor. A radiation oncologist manually contoured the target volume at the maximal inhalation phase of the respiratory cycle. An object constrained deformable image registration (DIR) method has been developed to track the target motion along the respiration at ten phases. The motion of the center of the target mass has been analyzed using the PCA to find out the principal motion components that were uncorrelated with each other. Two orthogonal image planes for cineMRI have been determined using this method to minimize the through plane motion during MRI based radiotherapy guidance. Results: 3D target respiratory motion for all 6 patients has been efficiently retrieved from 4DCT. In this process, the object constrained DIR demonstrated satisfactory accuracy and efficiency to enable the automatic motion tracking for clinical application. The average motion amplitude in the AP, lateral, and longitudinal directions were 3.6mm (min: 1.6mm, max: 5.6mm), 1.7mm (min: 0.6mm, max: 2.7mm), and 5.6mm (min: 1.8mm, max: 16.1mm), respectively. Based on PCA, the optimal orthogonal imaging planes were determined for cineMRI. The average angular difference between the PCA determined imaging planes and the traditional AP and lateral imaging planes were 47 and 31 degrees, respectively. After optimization, the average amplitude of through plane motion reduced from 3.6mm in AP images to 2.5mm (min:1.3mm, max:3.9mm); and from 1.7mm in lateral images to 0.6mm (min: 0.2mm, max:1.5mm), while the principal in plane motion amplitude increased from 5.6mm to 6.5mm (min: 2.8mm, max: 17mm). Conclusion: DIR and PCA can be used to optimize the orthogonal image planes of cineMRI to minimize the through plane motion during radiotherapy

  19. SU-G-BRA-03: PCA Based Imaging Angle Optimization for 2D Cine MRI Based Radiotherapy Guidance

    International Nuclear Information System (INIS)

    Chen, T; Yue, N; Jabbour, S; Zhang, M

    2016-01-01

    Purpose: To develop an imaging angle optimization methodology for orthogonal 2D cine MRI based radiotherapy guidance using Principal Component Analysis (PCA) of target motion retrieved from 4DCT. Methods: We retrospectively analyzed 4DCT of 6 patients with lung tumor. A radiation oncologist manually contoured the target volume at the maximal inhalation phase of the respiratory cycle. An object constrained deformable image registration (DIR) method has been developed to track the target motion along the respiration at ten phases. The motion of the center of the target mass has been analyzed using the PCA to find out the principal motion components that were uncorrelated with each other. Two orthogonal image planes for cineMRI have been determined using this method to minimize the through plane motion during MRI based radiotherapy guidance. Results: 3D target respiratory motion for all 6 patients has been efficiently retrieved from 4DCT. In this process, the object constrained DIR demonstrated satisfactory accuracy and efficiency to enable the automatic motion tracking for clinical application. The average motion amplitude in the AP, lateral, and longitudinal directions were 3.6mm (min: 1.6mm, max: 5.6mm), 1.7mm (min: 0.6mm, max: 2.7mm), and 5.6mm (min: 1.8mm, max: 16.1mm), respectively. Based on PCA, the optimal orthogonal imaging planes were determined for cineMRI. The average angular difference between the PCA determined imaging planes and the traditional AP and lateral imaging planes were 47 and 31 degrees, respectively. After optimization, the average amplitude of through plane motion reduced from 3.6mm in AP images to 2.5mm (min:1.3mm, max:3.9mm); and from 1.7mm in lateral images to 0.6mm (min: 0.2mm, max:1.5mm), while the principal in plane motion amplitude increased from 5.6mm to 6.5mm (min: 2.8mm, max: 17mm). Conclusion: DIR and PCA can be used to optimize the orthogonal image planes of cineMRI to minimize the through plane motion during radiotherapy

  20. Influence of Spatial Resolution in Three-dimensional Cine Phase Contrast Magnetic Resonance Imaging on the Accuracy of Hemodynamic Analysis.

    Science.gov (United States)

    Fukuyama, Atsushi; Isoda, Haruo; Morita, Kento; Mori, Marika; Watanabe, Tomoya; Ishiguro, Kenta; Komori, Yoshiaki; Kosugi, Takafumi

    2017-10-10

    We aim to elucidate the effect of spatial resolution of three-dimensional cine phase contrast magnetic resonance (3D cine PC MR) imaging on the accuracy of the blood flow analysis, and examine the optimal setting for spatial resolution using flow phantoms. The flow phantom has five types of acrylic pipes that represent human blood vessels (inner diameters: 15, 12, 9, 6, and 3 mm). The pipes were fixed with 1% agarose containing 0.025 mol/L gadolinium contrast agent. A blood-mimicking fluid with human blood property values was circulated through the pipes at a steady flow. Magnetic resonance (MR) images (three-directional phase images with speed information and magnitude images for information of shape) were acquired using the 3-Tesla MR system and receiving coil. Temporal changes in spatially-averaged velocity and maximum velocity were calculated using hemodynamic analysis software. We calculated the error rates of the flow velocities based on the volume flow rates measured with a flowmeter and examined measurement accuracy. When the acrylic pipe was the size of the thoracicoabdominal or cervical artery and the ratio of pixel size for the pipe was set at 30% or lower, spatially-averaged velocity measurements were highly accurate. When the pixel size ratio was set at 10% or lower, maximum velocity could be measured with high accuracy. It was difficult to accurately measure maximum velocity of the 3-mm pipe, which was the size of an intracranial major artery, but the error for spatially-averaged velocity was 20% or less. Flow velocity measurement accuracy of 3D cine PC MR imaging for pipes with inner sizes equivalent to vessels in the cervical and thoracicoabdominal arteries is good. The flow velocity accuracy for the pipe with a 3-mm-diameter that is equivalent to major intracranial arteries is poor for maximum velocity, but it is relatively good for spatially-averaged velocity.

  1. Impact of low signal intensity assessed by cine magnetic resonance imaging on detection of poorly viable myocardium in patients with prior myocardial infarction.

    Science.gov (United States)

    Ota, Shingo; Tanimoto, Takashi; Orii, Makoto; Hirata, Kumiko; Shiono, Yasutsugu; Shimamura, Kunihiro; Matsuo, Yoshiki; Yamano, Takashi; Ino, Yasushi; Kitabata, Hironori; Yamaguchi, Tomoyuki; Kubo, Takashi; Tanaka, Atsushi; Imanishi, Toshio; Akasaka, Takashi

    2015-05-13

    Late gadolinium enhancement magnetic resonance imaging (LGE-MRI) has been established as a modality to detect myocardial infarction (MI). However, the use of gadolinium contrast is limited in patients with advanced renal dysfunction. Although the signal intensity (SI) of infarct area assessed by cine MRI is low in some patients with prior MI, the prevalence and clinical significance of low SI has not been evaluated. The aim of this study was to evaluate how low SI assessed by cine MRI may relate to the myocardial viability in patients with prior MI. Fifty patients with prior MI underwent both cine MRI and LGE-MRI. The left ventricle was divided into 17 segments. The presence of low SI and the wall motion score (WMS) of each segment were assessed by cine MRI. The transmural extent of infarction was evaluated by LGE-MRI. LGE was detected in 329 of all 850 segments (39%). The low SI assessed by cine MRI was detected in 105 of 329 segments with LGE (32%). All segments with low SI had LGE. Of all 329 segments with LGE, the segments with low SI showed greater transmural extent of infarction (78 [72 - 84] % versus 53 [38 - 72] %, P cine MRI may be effective for detecting poorly viable myocardium in patients with prior MI.

  2. Compressed sensing cine imaging with high spatial or high temporal resolution for analysis of left ventricular function.

    Science.gov (United States)

    Goebel, Juliane; Nensa, Felix; Schemuth, Haemi P; Maderwald, Stefan; Gratz, Marcel; Quick, Harald H; Schlosser, Thomas; Nassenstein, Kai

    2016-08-01

    To assess two compressed sensing cine magnetic resonance imaging (MRI) sequences with high spatial or high temporal resolution in comparison to a reference steady-state free precession cine (SSFP) sequence for reliable quantification of left ventricular (LV) volumes. LV short axis stacks of two compressed sensing breath-hold cine sequences with high spatial resolution (SPARSE-SENSE HS: temporal resolution: 40 msec, in-plane resolution: 1.0 × 1.0 mm(2) ) and high temporal resolution (SPARSE-SENSE HT: temporal resolution: 11 msec, in-plane resolution: 1.7 × 1.7 mm(2) ) and of a reference cine SSFP sequence (standard SSFP: temporal resolution: 40 msec, in-plane resolution: 1.7 × 1.7 mm(2) ) were acquired in 16 healthy volunteers on a 1.5T MR system. LV parameters were analyzed semiautomatically twice by one reader and once by a second reader. The volumetric agreement between sequences was analyzed using paired t-test, Bland-Altman plots, and Passing-Bablock regression. Small differences were observed between standard SSFP and SPARSE-SENSE HS for stroke volume (SV; -7 ± 11 ml; P = 0.024), ejection fraction (EF; -2 ± 3%; P = 0.019), and myocardial mass (9 ± 9 g; P = 0.001), but not for end-diastolic volume (EDV; P = 0.079) and end-systolic volume (ESV; P = 0.266). No significant differences were observed between standard SSFP and SPARSE-SENSE HT regarding EDV (P = 0.956), SV (P = 0.088), and EF (P = 0.103), but for ESV (3 ± 5 ml; P = 0.039) and myocardial mass (8 ± 10 ml; P = 0.007). Bland-Altman analysis showed good agreement between the sequences (maximum bias ≤ -8%). Two compressed sensing cine sequences, one with high spatial resolution and one with high temporal resolution, showed good agreement with standard SSFP for LV volume assessment. J. Magn. Reson. Imaging 2016;44:366-374. © 2016 Wiley Periodicals, Inc.

  3. Monitoring tumor motion with on-line mega-voltage cone-beam computed tomography imaging in a cine mode

    International Nuclear Information System (INIS)

    Reitz, Bodo; Gayou, Olivier; Parda, David S; Miften, Moyed

    2008-01-01

    Accurate daily patient localization is becoming increasingly important in external-beam radiotherapy (RT). Mega-voltage cone-beam computed tomography (MV-CBCT) utilizing a therapy beam and an on-board electronic portal imager can be used to localize tumor volumes and verify the patient's position prior to treatment. MV-CBCT produces a static volumetric image and therefore can only account for inter-fractional changes. In this work, the feasibility of using the MV-CBCT raw data as a fluoroscopic series of portal images to monitor tumor changes due to e.g. respiratory motion was investigated. A method was developed to read and convert the CB raw data into a cine. To improve the contrast-to-noise ratio on the MV-CB projection data, image post-processing with filtering techniques was investigated. Volumes of interest from the planning CT were projected onto the MV-cine. Because of the small exposure and the varying thickness of the patient depending on the projection angle, soft-tissue contrast was limited. Tumor visibility as a function of tumor size and projection angle was studied. The method was well suited in the upper chest, where motion of the tumor as well as of the diaphragm could be clearly seen. In the cases of patients with non-small cell lung cancer with medium or large tumor masses, we verified that the tumor mass was always located within the PTV despite respiratory motion. However for small tumors the method is less applicable, because the visibility of those targets becomes marginal. Evaluation of motion in non-superior-inferior directions might also be limited for small tumor masses. Viewing MV-CBCT data in a cine mode adds to the utility of MV-CBCT for verification of tumor motion and for deriving individualized treatment margins

  4. Influence of longitudinal position on the evolution of steady-state signal in cardiac cine balanced steady-state free precession imaging.

    Science.gov (United States)

    Spear, Tyler J; Stromp, Tori A; Leung, Steve W; Vandsburger, Moriel H

    2017-11-01

    Emerging quantitative cardiac magnetic resonance imaging (CMRI) techniques use cine balanced steady-state free precession (bSSFP) to measure myocardial signal intensity and probe underlying physiological parameters. This correlation assumes that steady-state is maintained uniformly throughout the heart in space and time. To determine the effects of longitudinal cardiac motion and initial slice position on signal deviation in cine bSSFP imaging by comparing two-dimensional (2D) and three-dimensional (3D) acquisitions. Nine healthy volunteers completed cardiac MRI on a 1.5-T scanner. Short axis images were taken at six slice locations using both 2D and 3D cine bSSFP. 3D acquisitions spanned two slices above and below selected slice locations. Changes in myocardial signal intensity were measured across the cardiac cycle and compared to longitudinal shortening. For 2D cine bSSFP, 46% ± 9% of all frames and 84% ± 13% of end-diastolic frames remained within 10% of initial signal intensity. For 3D cine bSSFP the proportions increased to 87% ± 8% and 97% ± 5%. There was no correlation between longitudinal shortening and peak changes in myocardial signal. The initial slice position significantly impacted peak changes in signal intensity for 2D sequences ( P  cine bSSFP that is only restored at the center of a 3D excitation volume. During diastole, a transient steady-state is established similar to that achieved with 3D cine bSSFP regardless of slice location.

  5. Differentiation of recently infarcted myocardium from chronic myocardial scar: the value of contrast-enhanced SSFP-based cine MR imaging

    International Nuclear Information System (INIS)

    Kim, Kyoung Ah; Seo, Joon Beom; Do, Kyoung Hyun; Heo, Jeong Nam; Lee, Young Kyung; Song, Jae Woo; Lee, Jin Seong; Song, Koun Sik; Lim, Tae Hwan

    2006-01-01

    The purpose of this study is to demonstrate whether the signal intensity (SI) of myocardial infarction (MI) on contrast enhanced (CE)-cine MRI is useful for differentiating recently infarcted myocardium from chronic scar. This study included 24 patients with acute MI (36-84) years, mean age: 57) and 19 patients with chronic MI (44-80) years, mean age: 64). The diagnosis of acute MI was based on the presence of typical symptoms, i.e. elevation of the cardiac enzymes and the absence of any remote infarction history. The diagnosis of chronic MI was based on a history of MI or coronary artery disease of more than one month duration and on the absence of any recent MI within the previous six months. Retrospectively, the ECG-gated breath-hold cine imaging was performed in the short axis plane using a segmented, balanced, turbo-field, echo-pulse sequence two minutes after the administration of Gd-DTPA at a dose of 0.2 mmol/kg body weight. Delayed contrast-enhanced MRI (DCE MRI) in the same plane was performed 10 to 15 minutes after contrast administration, and this was served as the gold standard of reference. The SI of the infarcted myocardium on the CE-cine MRI was compared with that of the normal myocardium on the same image. The area of abnormal SI on the CE-cine MRI was compared with the area of hyper enhancement on the DCE MRI. The area of high SI on the CE-cine MRI was detected in 23 of 24 patients with acute MI (10 with homogenous high SI, 13 high SI with subendocardial low SI, and one with iso SI). The area of high SI on the CE- cine MRI was larger than that seen on the DCE MRI (ρ < 0.05). In contrast, the areas of chronic MI were seen as iso-SI with thin subendocardial low SI on the CE-cine MR in all the chronic MI patients. The presence of high SI on both the CE-cine MRI and the DCE MRI is more sensitive (95.8%) for determining the age of a MI than the presence of myocardial thinning (66.7%). The study showed the different SI patterns between recently

  6. Dynamic motion analysis of fetuses with central nervous system disorders by cine magnetic resonance imaging using fast imaging employing steady-state acquisition and parallel imaging: a preliminary result.

    Science.gov (United States)

    Guo, Wan-Yuo; Ono, Shigeki; Oi, Shizuo; Shen, Shu-Huei; Wong, Tai-Tong; Chung, Hsiao-Wen; Hung, Jeng-Hsiu

    2006-08-01

    The authors present a novel cine magnetic resonance (MR) imaging, two-dimensional (2D) fast imaging employing steady-state acquisition (FIESTA) technique with parallel imaging. It achieves temporal resolution at less than half a second as well as high spatial resolution cine imaging free of motion artifacts for evaluating the dynamic motion of fetuses in utero. The information obtained is used to predict postnatal outcome. Twenty-five fetuses with anomalies were studied. Ultrasonography demonstrated severe abnormalities in five of the fetuses; the other 20 fetuses constituted a control group. The cine fetal MR imaging demonstrated fetal head, neck, trunk, extremity, and finger as well as swallowing motions. Imaging findings were evaluated and compared in fetuses with major central nervous system (CNS) anomalies in five cases and minor CNS, non-CNS, or no anomalies in 20 cases. Normal motility was observed in the latter group. For fetuses in the former group, those with abnormal motility failed to survive after delivery, whereas those with normal motility survived with functioning preserved. The power deposition of radiofrequency, presented as specific absorption rate (SAR), was calculated. The SAR of FIESTA was approximately 13 times lower than that of conventional MR imaging of fetuses obtained using single-shot fast spin echo sequences. The following conclusions are drawn: 1) Fetal motion is no longer a limitation for prenatal imaging after the implementation of parallel imaging with 2D FIESTA, 2) Cine MR imaging illustrates fetal motion in utero with high clinical reliability, 3) For cases involving major CNS anomalies, cine MR imaging provides information on extremity motility in fetuses and serves as a prognostic indicator of postnatal outcome, and 4) The cine MR used to observe fetal activity is technically 2D and conceptually three-dimensional. It provides four-dimensional information for making proper and timely obstetrical and/or postnatal management

  7. The thoracic aortography by Gd-DTPA enhanced ultrafast cine MR imaging. Assessment of thoracic aortic dilatation in aging and in patients with hypertension and aortic valve disease

    International Nuclear Information System (INIS)

    Matsumura, Kentaro; Nakase, Emiko; Kawai, Ichiyoshi; Saito, Takayuki; Kikkawa, Nobutada; Haiyama, Toru

    1995-01-01

    To assess the morphology of thoracic aorta, we had a trial of Gd-DTPA enhanced ultrafast cine MR imaging on the thoracic aorta. This method was provided with high quality thoracic aortogram during 15-20 seconds. In patients without hypertension and aortic valve disease, dimensions of ascending aorta and aortic arch were significantly correlated with aging. In patients with hypertension, dimensions of ascending aorta and aortic arch were significantly dilated. In patients with aortic valve disease, thoracic aorta was diffusely enlarged, especially in ascending aorta. Gd-DTPA enhanced ultrafact cine MR imaging was useful to assess the thoracic aortic anatomy and diseases. (author)

  8. Spirometer-controlled Cine-Magnetic Resonance Imaging for Diagnosis of Tracheobronchomalacia in Pediatric Patients

    NARCIS (Netherlands)

    Ciet, P.; Wielopolski, P.; Manniesing, R.; Lever, S.; Bruijne, M. de; Morana, G.; Muzzio, P.C.; Lequin, M.H.; Tiddens, H.A.W.M.

    2014-01-01

    Tracheobronchomalacia (TBM) is defined as an excessive collapse of the intrathoracic trachea. Bronchoscopy is the gold standard to diagnose TBM, but bronchoscopy has major disadvantages, such as general anaesthesia. Cine-CT is a non-invasive alternative to diagnose TBM, but its use in children is

  9. TH-EF-BRA-08: A Novel Technique for Estimating Volumetric Cine MRI (VC-MRI) From Multi-Slice Sparsely Sampled Cine Images Using Motion Modeling and Free Form Deformation

    International Nuclear Information System (INIS)

    Harris, W; Yin, F; Wang, C; Chang, Z; Cai, J; Zhang, Y; Ren, L

    2016-01-01

    Purpose: To develop a technique to estimate on-board VC-MRI using multi-slice sparsely-sampled cine images, patient prior 4D-MRI, motion-modeling and free-form deformation for real-time 3D target verification of lung radiotherapy. Methods: A previous method has been developed to generate on-board VC-MRI by deforming prior MRI images based on a motion model(MM) extracted from prior 4D-MRI and a single-slice on-board 2D-cine image. In this study, free-form deformation(FD) was introduced to correct for errors in the MM when large anatomical changes exist. Multiple-slice sparsely-sampled on-board 2D-cine images located within the target are used to improve both the estimation accuracy and temporal resolution of VC-MRI. The on-board 2D-cine MRIs are acquired at 20–30frames/s by sampling only 10% of the k-space on Cartesian grid, with 85% of that taken at the central k-space. The method was evaluated using XCAT(computerized patient model) simulation of lung cancer patients with various anatomical and respirational changes from prior 4D-MRI to onboard volume. The accuracy was evaluated using Volume-Percent-Difference(VPD) and Center-of-Mass-Shift(COMS) of the estimated tumor volume. Effects of region-of-interest(ROI) selection, 2D-cine slice orientation, slice number and slice location on the estimation accuracy were evaluated. Results: VCMRI estimated using 10 sparsely-sampled sagittal 2D-cine MRIs achieved VPD/COMS of 9.07±3.54%/0.45±0.53mm among all scenarios based on estimation with ROI_MM-ROI_FD. The FD optimization improved estimation significantly for scenarios with anatomical changes. Using ROI-FD achieved better estimation than global-FD. Changing the multi-slice orientation to axial, coronal, and axial/sagittal orthogonal reduced the accuracy of VCMRI to VPD/COMS of 19.47±15.74%/1.57±2.54mm, 20.70±9.97%/2.34±0.92mm, and 16.02±13.79%/0.60±0.82mm, respectively. Reducing the number of cines to 8 enhanced temporal resolution of VC-MRI by 25% while

  10. TH-EF-BRA-08: A Novel Technique for Estimating Volumetric Cine MRI (VC-MRI) From Multi-Slice Sparsely Sampled Cine Images Using Motion Modeling and Free Form Deformation

    Energy Technology Data Exchange (ETDEWEB)

    Harris, W; Yin, F; Wang, C; Chang, Z; Cai, J; Zhang, Y; Ren, L [Duke University Medical Center, Durham, NC (United States)

    2016-06-15

    Purpose: To develop a technique to estimate on-board VC-MRI using multi-slice sparsely-sampled cine images, patient prior 4D-MRI, motion-modeling and free-form deformation for real-time 3D target verification of lung radiotherapy. Methods: A previous method has been developed to generate on-board VC-MRI by deforming prior MRI images based on a motion model(MM) extracted from prior 4D-MRI and a single-slice on-board 2D-cine image. In this study, free-form deformation(FD) was introduced to correct for errors in the MM when large anatomical changes exist. Multiple-slice sparsely-sampled on-board 2D-cine images located within the target are used to improve both the estimation accuracy and temporal resolution of VC-MRI. The on-board 2D-cine MRIs are acquired at 20–30frames/s by sampling only 10% of the k-space on Cartesian grid, with 85% of that taken at the central k-space. The method was evaluated using XCAT(computerized patient model) simulation of lung cancer patients with various anatomical and respirational changes from prior 4D-MRI to onboard volume. The accuracy was evaluated using Volume-Percent-Difference(VPD) and Center-of-Mass-Shift(COMS) of the estimated tumor volume. Effects of region-of-interest(ROI) selection, 2D-cine slice orientation, slice number and slice location on the estimation accuracy were evaluated. Results: VCMRI estimated using 10 sparsely-sampled sagittal 2D-cine MRIs achieved VPD/COMS of 9.07±3.54%/0.45±0.53mm among all scenarios based on estimation with ROI-MM-ROI-FD. The FD optimization improved estimation significantly for scenarios with anatomical changes. Using ROI-FD achieved better estimation than global-FD. Changing the multi-slice orientation to axial, coronal, and axial/sagittal orthogonal reduced the accuracy of VCMRI to VPD/COMS of 19.47±15.74%/1.57±2.54mm, 20.70±9.97%/2.34±0.92mm, and 16.02±13.79%/0.60±0.82mm, respectively. Reducing the number of cines to 8 enhanced temporal resolution of VC-MRI by 25% while

  11. Strain analysis in CRT candidates using the novel segment length in cine (SLICE) post-processing technique on standard CMR cine images

    NARCIS (Netherlands)

    Zweerink, A.; Allaart, C.P.; Kuijer, J.P.A.; Wu, L.; Beek, A.M.; Ven, P.M. van de; Meine, M.; Croisille, P.; Clarysse, P.; Rossum, A.C. van; Nijveldt, R.

    2017-01-01

    OBJECTIVES: Although myocardial strain analysis is a potential tool to improve patient selection for cardiac resynchronization therapy (CRT), there is currently no validated clinical approach to derive segmental strains. We evaluated the novel segment length in cine (SLICE) technique to derive

  12. Use of the statistical control of processes through checking before treatment realised with the ionization chamber and by electronic portal system of imaging (E.P.I.D.) in intensity modulated radiotherapy (I.M.R.T.); Utilisation de la maitrise statistique des processus dans le cadre des controles avant traitement realises avec la chambre d'ionisation et par systeme d'imagerie portale electronique (EPID) en radiotherapie conformationnelle avec modulation d'intensite (RCMI)

    Energy Technology Data Exchange (ETDEWEB)

    Villani, N.; Gerard, K.; Noel, A. [Nancy univ., Lab. de Recherche en Radiophysique, CRAN UMR 7039, CNRS, Centre Alexis-Vautrin, 54 - Vandoeuvre-les-Nancy (France); Marchesi, V.; Huger, S. [Centre Alexis-Vautrin, Unite de Radiophysique Medicale, 54 - Vandoeuvre-les-Nancy (France)

    2009-10-15

    The expected results are to demonstrate that it is possible to reduce the times devoted to the pre-treatment controls, while keeping an optimal safety, on substituting the measures of the ionization chamber by this one of the electronic portal imaging device (E.P.I.D.). (N.C.)

  13. Portal hypertension in patients with cirrhosis: indirect assessment of hepatic venous pressure gradient by measuring azygos flow with 2D-cine phase-contrast magnetic resonance imaging.

    Science.gov (United States)

    Gouya, Hervé; Grabar, Sophie; Vignaux, Olivier; Saade, Anastasia; Pol, Stanislas; Legmann, Paul; Sogni, Philippe

    2016-07-01

    To measure azygos, portal and aortic flow by two-dimensional cine phase-contrast magnetic resonance imaging (2D-cine PC MRI), and to compare the MRI values to hepatic venous pressure gradient (HVPG) measurements, in patients with cirrhosis. Sixty-nine patients with cirrhosis were prospectively included. All patients underwent HVPG measurements, upper gastrointestinal endoscopy and 2D-cine PC MRI measurements of azygos, portal and aortic blood flow. Univariate and multivariate regression analyses were used to evaluate the correlation between the blood flow and HVPG. The performance of 2D-cine PC MRI to diagnose severe portal hypertension (HVPG ≥ 16 mmHg) was determined by receiver operating characteristic curve (ROC) analysis, and area under the curves (AUC) were compared. Azygos and aortic flow values were associated with HVPG in univariate linear regression model. Azygos flow (p portal blood flow (AUC = 0.40 (95 % CI [0.25-0.54]). 2D-cine PC MRI is a promising technique to evaluate significant portal hypertension in patients with cirrhosis. • Noninvasive HVPG assessment can be performed with MRI azygos flow. • Azygos MRI flow is an easy-to-measure marker to detect significant portal hypertension. • MRI flow is more specific that varice grade to detect portal hypertension.

  14. Impact of Surgical Evaluation of Additional Cine Magnetic Resonance Imaging for Advanced Thymoma with Infiltration of Adjacent Structures: The Thoracic Surgeon's View.

    Science.gov (United States)

    Ried, Michael; Hnevkovsky, Stefanie; Neu, Reiner; von Süßkind-Schwendi, Marietta; Götz, Andrea; Hamer, Okka W; Schalke, Berthold; Hofmann, Hans-Stefan

    2017-04-01

    Background  Preoperative radiological assessment is important for clarification of surgical operability for advanced thymic tumors. Objective was to determine the feasibility of magnetic resonance imaging (MRI) with cine sequences for evaluation of cardiovascular tumor invasion. Patients and Methods  This prospective study included patients with advanced thymoma, who underwent surgical resection. All patients received preoperative computed tomography (CT) scan and cine MRI. Results  Tumor infiltration was surgically confirmed in the pericardium ( n  = 12), myocardium ( n  = 1), superior caval vein (SCV; n  = 3), and aorta ( n  = 2). A macroscopic complete resection was possible in 10 patients, whereas 2 patients with aortic or myocardial tumor invasion had R2 resection. The positive predictive value (PPV) was 50% for cine MRI compared with 0% for CT scan regarding myocardial tumor infiltration. The PPV for tumor infiltration of the aorta was 50%, with a higher sensitivity for the CT scan (100 vs. 50%). Infiltration of the SCV could be detected slightly better with cine MRI (PPV 75 vs. 66.7%). Conclusion  Cine MRI seems to improve the accuracy of preoperative staging of advanced thymoma regarding infiltration of cardiovascular structures and supports the surgical approach. Georg Thieme Verlag KG Stuttgart · New York.

  15. Portal hypertension in patients with cirrhosis: indirect assessment of hepatic venous pressure gradient by measuring azygos flow with 2D-cine phase-contrast magnetic resonance imaging

    International Nuclear Information System (INIS)

    Gouya, Herve; Vignaux, Olivier; Saade, Anastasia; Legmann, Paul; Grabar, Sophie; Pol, Stanislas; Sogni, Philippe

    2016-01-01

    To measure azygos, portal and aortic flow by two-dimensional cine phase-contrast magnetic resonance imaging (2D-cine PC MRI), and to compare the MRI values to hepatic venous pressure gradient (HVPG) measurements, in patients with cirrhosis. Sixty-nine patients with cirrhosis were prospectively included. All patients underwent HVPG measurements, upper gastrointestinal endoscopy and 2D-cine PC MRI measurements of azygos, portal and aortic blood flow. Univariate and multivariate regression analyses were used to evaluate the correlation between the blood flow and HVPG. The performance of 2D-cine PC MRI to diagnose severe portal hypertension (HVPG ≥ 16 mmHg) was determined by receiver operating characteristic curve (ROC) analysis, and area under the curves (AUC) were compared. Azygos and aortic flow values were associated with HVPG in univariate linear regression model. Azygos flow (p < 10 -3 ), aortic flow (p = 0.001), age (p = 0.001) and presence of varices (p < 10 -3 ) were independently associated with HVPG. Azygos flow (AUC = 0.96 (95 % CI) [0.91-1.00]) had significantly higher AUC than aortic (AUC = 0.64 (95 % CI) [0.51-0.77]) or portal blood flow (AUC = 0.40 (95 % CI) [0.25-0.54]). 2D-cine PC MRI is a promising technique to evaluate significant portal hypertension in patients with cirrhosis. (orig.)

  16. Fast cine-magnetic resonance imaging point tracking for prostate cancer radiation therapy planning

    International Nuclear Information System (INIS)

    Dowling, J; Chandra, S; Dang, K; Fox, Chris D; Gill, Suki; Kron, T; Pham, D; Foroudi, F

    2014-01-01

    The analysis of intra-fraction organ motion is important for improving the precision of radiation therapy treatment delivery. One method to quantify this motion is for one or more observers to manually identify anatomic points of interest (POIs) on each slice of a cine-MRI sequence. However this is labour intensive and inter- and intra- observer variation can introduce uncertainty. In this paper a fast method for non-rigid registration based point tracking in cine-MRI sagittal and coronal series is described which identifies POIs in 0.98 seconds per sagittal slice and 1.35 seconds per coronal slice. The manual and automatic points were highly correlated (r>0.99, p<0.001) for all organs and the difference generally less than 1mm. For prostate planning peristalsis and rectal gas can result in unpredictable out of plane motion, suggesting the results may require manual verification.

  17. Real-time cardiac magnetic resonance cine imaging with sparse sampling and iterative reconstruction for left-ventricular measures: Comparison with gold-standard segmented steady-state free precession.

    Science.gov (United States)

    Camargo, Gabriel C; Erthal, Fernanda; Sabioni, Leticia; Penna, Filipe; Strecker, Ralph; Schmidt, Michaela; Zenge, Michael O; Lima, Ronaldo de S L; Gottlieb, Ilan

    2017-05-01

    Segmented cine imaging with a steady-state free-precession sequence (Cine-SSFP) is currently the gold standard technique for measuring ventricular volumes and mass, but due to multi breath-hold (BH) requirements, it is prone to misalignment of consecutive slices, time consuming and dependent on respiratory capacity. Real-time cine avoids those limitations, but poor spatial and temporal resolution of conventional sequences has prevented its routine application. We sought to examine the accuracy and feasibility of a newly developed real-time sequence with aggressive under-sampling of k-space using sparse sampling and iterative reconstruction (Cine-RT). Stacks of short-axis cines were acquired covering both ventricles in a 1.5T system using gold standard Cine-SSFP and Cine-RT. Acquisition parameters for Cine-SSFP were: acquisition matrix of 224×196, temporal resolution of 39ms, retrospective gating, with an average of 8 heartbeats per slice and 1-2 slices/BH. For Cine-RT: acquisition matrix of 224×196, sparse sampling net acceleration factor of 11.3, temporal resolution of 41ms, prospective gating, real-time acquisition of 1 heart-beat/slice and all slices in one BH. LV contours were drawn at end diastole and systole to derive LV volumes and mass. Forty-one consecutive patients (15 male; 41±17years) in sinus rhythm were successfully included. All images from Cine-SSFP and Cine-RT were considered to have excellent quality. Cine-RT-derived LV volumes and mass were slightly underestimated but strongly correlated with gold standard Cine-SSFP. Inter- and intra-observer analysis presented similar results between both sequences. Cine-RT featuring sparse sampling and iterative reconstruction can achieve spatial and temporal resolution equivalent to Cine-SSFP, providing excellent image quality, with similar precision measurements and highly correlated and only slightly underestimated volume and mass values. Copyright © 2017 Elsevier Inc. All rights reserved.

  18. Self-gated CINE MRI for combined contrast-enhanced imaging and wall-stiffness measurements of murine aortic atherosclerotic lesions

    NARCIS (Netherlands)

    den Adel, Brigit; van der Graaf, Linda M.; Strijkers, Gustav J.; Lamb, Hildo J.; Poelmann, Robert E.; van der Weerd, Louise

    2013-01-01

    High-resolution contrast-enhanced imaging of the murine atherosclerotic vessel wall is difficult due to unpredictable flow artifacts, motion of the thin artery wall and problems with flow suppression in the presence of a circulating contrast agent. We applied a 2D-FLASH retrospective-gated CINE MRI

  19. A novel diagnostic aid for intra-abdominal adhesion detection in cine-MR imaging: Pilot study and initial diagnostic impressions.

    Science.gov (United States)

    Randall, David; Joosten, Frank; ten Broek, Richard; Gillott, Richard; Bardhan, Karna Dev; Strik, Chema; Prins, Wiesje; van Goor, Harry; Fenner, John

    2017-07-14

    A non-invasive diagnostic technique for abdominal adhesions is not currently available. Capture of abdominal motion due to respiration in cine-MRI has shown promise, but is difficult to interpret. This article explores the value of a complimentary diagnostic aid to facilitate the non-invasive detection of abdominal adhesions using cine-MRI. An image processing technique was developed to quantify the amount of sliding that occurs between the organs of the abdomen and the abdominal wall in sagittal cine-MRI slices. The technique produces a 'sheargram' which depicts the amount of sliding which has occurred over 1-3 respiratory cycles. A retrospective cohort of 52 patients, scanned for suspected adhesions, made 281 cine-MRI sagittal slices available for processing. The resulting sheargrams were reported by two operators and compared to expert clinical judgement of the cine-MRI scans. The sheargram matched clinical judgement in 84% of all sagittal slices and 93-96% of positive adhesions were identified on the sheargram. The sheargram displayed a slight skew towards sensitivity over specificity, with a high positive adhesion detection rate but at the expense of false positives. Good correlation between sheargram and absence/presence of inferred adhesions indicates quantification of sliding motion has potential to aid adhesion detection in cine-MRI. Advances in Knowledge: This is the first attempt to clinically evaluate a novel image processing technique quantifying the sliding motion of the abdominal contents against the abdominal wall. The results of this pilot study reveal its potential as a diagnostic aid for detection of abdominal adhesions.

  20. Dose-response characteristics of an amorphous silicon EPID

    International Nuclear Information System (INIS)

    Winkler, Peter; Hefner, Alfred; Georg, Dietmar

    2005-01-01

    Electronic portal imaging devices (EPIDs) were originally developed for the purpose of patient setup verification. Nowadays, they are increasingly used as dosimeters (e.g., for IMRT verification and linac-specific QA). A prerequisite for any clinical dosimetric application is a detailed understanding of the detector's dose-response behavior. The aim of this study is to investigate the dosimetric properties of an amorphous silicon EPID (Elekta IVIEWGT) with respect to three photon beam qualities: 6, 10, and 25 MV. The EPID showed an excellent temporal stability on short term as well as on long term scales. The stability throughout the day was strongly influenced by warming up, which took several hours and affected EPID response by 2.5%. Ghosting effects increased the sensitivity of the EPID. They became more pronounced with decreasing time intervals between two exposures as well as with increasing dose. Due to ghosting, changes in pixel sensitivity amounted up to 16% (locally) for the 25 MV photon beam. It was observed that the response characteristics of our EPID depended on dose as well as on dose rate. Doubling the dose rate increased the EPID sensitivity by 1.5%. This behavior was successfully attributed to a dose per frame effect, i.e., a nonlinear relationship between the EPID signal and the dose which was delivered to the panel between two successive readouts. The sensitivity was found to vary up to 10% in the range of 1 to 1000 monitor units. This variation was governed by two independent effects. For low doses, the EPID signal was reduced due to the linac's changing dose rate during startup. Furthermore, the detector reading was influenced by intrabeam variations of EPID sensitivity, namely, an increase of detector response during uniform exposure. For the beam qualities which were used, the response characteristics of the EPID did not depend on energy. Differences in relative dose-response curves resulted from energy dependent temporal output

  1. 2D XD-GRASP provides better image quality than conventional 2D cardiac cine MRI for patients who cannot suspend respiration

    Science.gov (United States)

    Piekarski, Eve; Chitiboi, Teodora; Ramb, Rebecca; Latson, Larry A; Bhatla, Puneet; Feng, Li; Axel, Leon

    2017-01-01

    Object Residual respiratory motion degrades image quality in conventional cardiac cine MRI (CCMR). We evaluated whether a free-breathing (FB) radial imaging CCMR sequence with compressed sensing reconstruction (eXtra-Dimension (e.g. cardiac and respiratory phases) Golden-angle RAdial Sparse Parallel, or XD-GRASP) could provide better image quality than a conventional Cartesian breath-held (BH) sequence, in an unselected population of patients undergoing clinical CCMR. Material and Methods 101 patients who underwent BH and FB imaging in a mid-ventricular short-axis plane at a matching location were included. Visual and quantitative image analysis was performed by two blinded experienced readers, using a 5-point qualitative scale to score overall image quality and visual signal-to-noise ratio (SNR) grade, with measures of noise and sharpness. End-diastole (ED) and end-systole (ES) left-ventricular areas were also measured and compared for both BH and FB images. Results Image quality was generally better with the BH cines (overall quality grade BH vs FB: 4 vs 2.9, pXD-GRASP CCMR was visually inferior to conventional BH cardiac cine in general, it provided improved image quality in the subgroup of patients presenting respiratory motion-induced artifacts on breath-held images. PMID:29067539

  2. Cardiac MRI: evaluation of phonocardiogram-gated cine imaging for the assessment of global und regional left ventricular function in clinical routine

    International Nuclear Information System (INIS)

    Nassenstein, Kai; Schlosser, Thomas; Orzada, Stephan; Haering, Lars; Czylwik, Andreas; Zenge, Michael; Mueller, Edgar; Eberle, Holger; Bruder, Oliver; Ladd, Mark E.; Maderwald, Stefan

    2012-01-01

    To validate a phonocardiogram (PCG)-gated cine imaging approach for the assessment of left ventricular (LV) function. In this prospective study, cine MR imaging of the LV was performed twice in 79 patients by using retrospectively PCG- and retrospectively ECG-gated cine SSFP sequences at 1.5 T. End-diastolic volumes (EDV), end-systolic volumes (ESV), stroke volumes (SV), ejection fraction (EF), muscle mass (MM), as well as regional wall motion were assessed. Subgroup analyses were performed for patients with valvular defects and for patients with dysrhythmia. PCG-gated imaging was feasible in 75 (95%) patients, ECG-gating in all patients. Excellent correlations were observed for all volumetric parameters (r > 0.98 for all variables analysed). No significant differences were observed for EDV (-0.24 ± 3.14 mL, P = 0.5133), ESV (-0.04 ± 2.36 mL, P = 0.8951), SV (-0.20 ± 3.41 mL, P = 0.6083), EF (-0.16 ± 1.98%, P = 0.4910), or MM (0.31 ± 4.2 g, P = 0.7067) for the entire study cohort, nor for either of the subgroups. PCG- and ECG-gated cine imaging revealed similar results for regional wall motion analyses (115 vs. 119 segments with wall motion abnormalities, P = 0.3652). The present study demonstrates that PCG-gated cine imaging enables accurate assessment of global and regional LV function in the vast majority of patients in clinical routine. (orig.)

  3. TH-CD-207A-02: Implementation of Live EPID-Based Inspiration Level Assessment (LEILA) for Deepinspiration Breath-Hold (DIBH) Monitoring Using MV Fluoroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Lehmann, J [Calvary Mater Newcastle, Newcastle (Australia); The University of Sydney, Sydney (Australia); The University of Newcastle, Newcastle (Australia); Sun, J; Fuangrod, T; Bhatia, S [Calvary Mater Newcastle, Newcastle (Australia); Doebrich, M; Greer, P [Calvary Mater Newcastle, Newcastle (Australia); The University of Newcastle, Newcastle (Australia); Zwan, B [The University of Newcastle, Newcastle (Australia); Central Coast Cancer Centre, Gosford (Australia)

    2016-06-15

    Purpose: As prior work has shown that current DIBH monitoring approaches using surrogate measures (marker block on chest) do not always correspond with the clinical quantity of interest (lung depth, LD), a software tool and workflow are introduced to use MV fluoroscopy during treatment for real-time / Live EPID-based Inspiration Level Assessment (LEILA). Methods: A prototype software tool calculates and displays the LD during the treatment of left sided breast cancer. Calculations are based on MV cine images which are acquired with the treatment beam thereby not incurring any additional imaging dose. Image capture and processing are implemented using a dedicated frame grabber computer. The calculation engine automatically detects image orientation and includes provisions for large treatment fields that exceed the size of the EPID panel. LD is measured along a line profile in the middle of the field. LEILA’s interface displays the current MV image, a reference image (DRR), the current LD, as well as a trace of LD over treatment time. The display includes patient specific LD tolerances. Tolerances are specified for each field and loaded before the treatment. A visual warning is generated when the tolerance is exceeded. LEILA is initially run in parallel with current DIBH techniques. When later run by itself DIBH setup will be done using skin marks and room laser. Results: Offline tests of LEILA confirmed accurate automatic LD measurement for a variety of patient geometries. Deployment of the EPID during all left sided breast treatments was well tolerated by patients and staff during a multi-month pilot. The frame grabber provides 11 frames-per-second; the MATLAB based LEILA prototype software can analyze five frames-per-second standalone on standard desktop hardware. Conclusion: LEILA provides an automated approach to quantitatively monitor LD on MV images during DIBH treatment. Future improvements include a database and further speed optimization.

  4. Development and clinical evaluation of automatic fiducial detection for tumor tracking in cine megavoltage images during volumetric modulated arc therapy

    International Nuclear Information System (INIS)

    Azcona, Juan Diego; Li Ruijiang; Mok, Edward; Hancock, Steven; Xing Lei

    2013-01-01

    Purpose: Real-time tracking of implanted fiducials in cine megavoltage (MV) imaging during volumetric modulated arc therapy (VMAT) delivery is complicated due to the inherent low contrast of MV images and potential blockage of dynamic leaves configurations. The purpose of this work is to develop a clinically practical autodetection algorithm for motion management during VMAT. Methods: The expected field-specific segments and the planned fiducial position from the Eclipse (Varian Medical Systems, Palo Alto, CA) treatment planning system were projected onto the MV images. The fiducials were enhanced by applying a Laplacian of Gaussian filter in the spatial domain for each image, with a blob-shaped object as the impulse response. The search of implanted fiducials was then performed on a region of interest centered on the projection of the fiducial when it was within an open field including the case when it was close to the field edge or partially occluded by the leaves. A universal template formula was proposed for template matching and normalized cross correlation was employed for its simplicity and computational efficiency. The search region for every image was adaptively updated through a prediction model that employed the 3D position of the fiducial estimated from the localized positions in previous images. This prediction model allowed the actual fiducial position to be tracked dynamically and was used to initialize the search region. The artifacts caused by electronic interference during the acquisition were effectively removed. A score map was computed by combining both morphological information and image intensity. The pixel location with the highest score was selected as the detected fiducial position. The sets of cine MV images taken during treatment were analyzed with in-house developed software written in MATLAB (The Mathworks, Inc., Natick, MA). Five prostate patients were analyzed to assess the algorithm performance by measuring their positioning

  5. Measurement of blood flow in the left ventricle and aorta using clinical 2D cine phase-contrast magnetic resonance imaging

    International Nuclear Information System (INIS)

    Nakamura, Masanori; Wada, Shigeo; Yokosawa, Suguru; Yamaguchi, Takami; Isoda, Haruo; Takeda, Hiroyasu

    2007-01-01

    A recent development in phase-contrast magnetic resonance imaging (cine PC-MRI) allows the detailed measurement of the blood flow in humans. The objectives of this study are twofold: to discuss the utility of clinical two-dimensional (2D) cine PC-MRI as a practical tool for analyzing hemodynamics in the aorta and left ventricle; to provide flow information at those places as references for computational fluid dynamics studies. Using 2D cine PC-MRI, we mapped velocity profiles at various cross sections of the aorta and left ventricle. The results illustrated the main flow events in the left ventricle during the cardiac cycle, such as ventricular ejection and suction, while the secondary flows were less clear. The velocity profile at the entrance of the ascending aorta appeared to be slightly skewed posteriorly in early systole, but the flow in the central zone of the section was rapid. The estimated stroke volume, peak Reynolds numbers, and Womersley numbers were within the normal physiological range. A sequence of secondary flow images from the plane of the aortic valve to the descending aorta revealed the evolution of a helical flow within the aorta. Flows entering the aortic branches were captured well. Those results demonstrate that clinical 2D cine PC-MRI is a practical adjunct for analyzing blood flow in vivo and would be useful as references to check validity of flow dynamics obtained by computer simulations. (author)

  6. Significance of spin-echo intracardiac signal during cine cardiac MR imaging

    International Nuclear Information System (INIS)

    Feiglin, D.H.I.; O'Donnell, J.K.

    1987-01-01

    Thirty patient studies were performed using several multisection spin multi-echo pulse sequences (SEPS) formattable into the cine mode, with repetition time (TR)≤RR interval and 18 msec ≤ echo time (TE) ≤ 64 msec. Thirteen studies were performed in patients with various cardiomyopathies, ten in patients with cardiac tumors, and seven in healthy volunteers. The SEPS in the multi-echo acquisition format differentiated between tumor and stasis in terms of signal behavior. Healthy subjects may exhibit stasis of flow adjacent to the endocardium during the cardiac cycle

  7. Hemodynamic study on flow patterns in the carotid bifurcation before and after carotid endarterectomy using cine magnetic resonance imaging

    Energy Technology Data Exchange (ETDEWEB)

    Yamane, Kanji; Shima, Takeshi; Okada, Yoshikazu; Nishida, Masahiro; Okita, Shinji; Hanaguri, Katsuro [Chugoku Rousai Hospital, Kure, Hiroshima (Japan)

    1993-11-01

    Blood flow in the cervical carotid bifurcation was investigated by cine magnetic resonance imaging. In patients with stenosis, a low-intensity stream was demonstrated from the beginning of the carotid bulb, which was more distinct in the systolic phase. In patients with stenotic carotid bifurcations,the low-intensity flow was also present but was more prominent than in the non-stenotic bifurcation. This low-intensity stream may be due to the change from steady to turbulent flow due to the geometric characteristics of the carotid bifurcation or atheromatous plaque, similar to the flow separation phenomenon in fluid dynamics because of the coincidence of location and flow pattern. After carotid endarterectomy, turbulent flow was seen at the proximal and distal ends of the endarterectomy. Close follow-up and administration of antiplatelet agents are necessary to prevent restenosis due to mural thrombosis induced by such turbulent flow. (author).

  8. Identification and Assessment of Paradoxical Ventricular Wall Motion Using ECG Gated Blood Pool Scan - Comparison of cine Loop , Phase Analysis and Paradox Image -

    International Nuclear Information System (INIS)

    Lee, Jae Tae; Kim, Gwang Weon; Lee, Kyu Bo; Chung, Byung Chun; Whang, Kee Suk; Chae, Sung Chul; Paek, Wee Hyun; Cheon, Jae Eun; Lee, Hyong Woo; Chung, Jin Hong

    1990-01-01

    Sixty-four patients with paradoxical ventricular wall motion noticed both in angiocardiography or 2-dimensional echocardiography were assessed by ECG gated blood pool scan (GBPS). Endless cine loop image, phase and amplitude images and paradox image obtained by visual inspection of each cardiac beat or Fourier transformation of acquired raw data were investigated to determine the incremental value of GBPS with these processing methods for identification of paradoxical ventricular wall motion. The results were as follows:1) Paradoxical wall motions were observed on interventricular septum in 34 cases, left ventricular free wall in 26 and right ventricular wall in 24. Underlying heart diseases were is chemic (23 cases) valvular(9), congenital heart disease (12), cardiomyopathy (5), pericardial effusion(5), post cardiac surgery(3), corpulmonale (2), endocarditis (l) and right ventricular tumor(l). 2) Left ventricular ejection fractions of patients with paradoxical left ventricular wall motion were significantly lower than those with paradoxical septal motion (p <0.005). 3) The sensitivity of each processing methods for detecting paradoxical wall motion was 76.9% by phase analysis, 74.6% by endless cine loop mapping and 68.4% by paradox image manipulation respectively. Paradoxial motions visualized only in phase, paradox or both images were appeared as hypokinesia or akinesia in cine loop image. 4) All events could be identified by at least one of above three processing methods, however only 34 cases (48.4%) showed the paradoxical motions in all of the three images. By these findings, we concluded that simultaneous inspection of all above three processing methods-endless cine loop, phase analysis and paradox image is necessary for accurate identification and assessment of paradoxical ventricular wall motion when performing GBPS.

  9. Identification and Assessment of Paradoxical Ventricular Wall Motion Using ECG Gated Blood Pool Scan - Comparison of cine Loop , Phase Analysis and Paradox Image -

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Jae Tae; Kim, Gwang Weon; Lee, Kyu Bo; Chung, Byung Chun; Whang, Kee Suk; Chae, Sung Chul; Paek, Wee Hyun; Cheon, Jae Eun [Kyungpook National University School of Medicine, Daegu (Korea, Republic of); Lee, Hyong Woo; Chung, Jin Hong [Yeongnam National University College of Medicine, Daegu (Korea, Republic of)

    1990-07-15

    Sixty-four patients with paradoxical ventricular wall motion noticed both in angiocardiography or 2-dimensional echocardiography were assessed by ECG gated blood pool scan (GBPS). Endless cine loop image, phase and amplitude images and paradox image obtained by visual inspection of each cardiac beat or Fourier transformation of acquired raw data were investigated to determine the incremental value of GBPS with these processing methods for identification of paradoxical ventricular wall motion. The results were as follows:1) Paradoxical wall motions were observed on interventricular septum in 34 cases, left ventricular free wall in 26 and right ventricular wall in 24. Underlying heart diseases were is chemic (23 cases) valvular(9), congenital heart disease (12), cardiomyopathy (5), pericardial effusion(5), post cardiac surgery(3), corpulmonale (2), endocarditis (l) and right ventricular tumor(l). 2) Left ventricular ejection fractions of patients with paradoxical left ventricular wall motion were significantly lower than those with paradoxical septal motion (p <0.005). 3) The sensitivity of each processing methods for detecting paradoxical wall motion was 76.9% by phase analysis, 74.6% by endless cine loop mapping and 68.4% by paradox image manipulation respectively. Paradoxial motions visualized only in phase, paradox or both images were appeared as hypokinesia or akinesia in cine loop image. 4) All events could be identified by at least one of above three processing methods, however only 34 cases (48.4%) showed the paradoxical motions in all of the three images. By these findings, we concluded that simultaneous inspection of all above three processing methods-endless cine loop, phase analysis and paradox image is necessary for accurate identification and assessment of paradoxical ventricular wall motion when performing GBPS.

  10. Analysis of blood flow patterns in aortic aneurysm by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Matsuoka, Hiroshi

    1993-01-01

    Cine MRI (0.5 T) using rephased gradient echo technique was performed to study the patterns of blood flow in the aortic aneurysm of 16 patients with aortic aneurysm, and the data were compared with those of 5 healthy volunteers. In the transaxial section, the blood flow in normal aorta appeared as homogeneous high intensity during systole. On the other hand, the blood flow in the aneurysm appeared as inhomogeneous flow enhancement with flow void. In the sagittal scan, the homogeneous flow enhancement in a normal aorta was also observed during systole and its apex of flow enhancement was 'taper'. The blood flow patterns in the aneurysm were classified as 'irregular', 'zonal', 'eddy', and 'obscure' depending on the contrast of flow enhancement and flow void. Their apexes were 'taper' or 'round'. The blood flow patterns in the aneurysm were related to the size of aneurysm. In patients with a large size 'aneurysm, their flow patterns were 'eddy' or 'obscure' and the flow enhancement was 'round'. On the other hand, in patients with a small size aneurysm, their flow patterns were 'irregular' or 'zonal', and their flow enhancement was 'taper'. Though the exact mechanism of abnormal flow patterns in an aortic aneurysm remains to be determined, cine MRI gives helpful informations in assessing blood flow dynamics in the aneurysm. (author)

  11. Differentiation between acute and chronic myocardial infarction by means of texture analysis of late gadolinium enhancement and cine cardiac magnetic resonance imaging.

    Science.gov (United States)

    Larroza, Andrés; Materka, Andrzej; López-Lereu, María P; Monmeneu, José V; Bodí, Vicente; Moratal, David

    2017-07-01

    The purpose of this study was to differentiate acute from chronic myocardial infarction using machine learning techniques and texture features extracted from cardiac magnetic resonance imaging (MRI). The study group comprised 22 cases with acute myocardial infarction (AMI) and 22 cases with chronic myocardial infarction (CMI). Cine and late gadolinium enhancement (LGE) MRI were analyzed independently to differentiate AMI from CMI. A total of 279 texture features were extracted from predefined regions of interest (ROIs): the infarcted area on LGE MRI, and the entire myocardium on cine MRI. Classification performance was evaluated by a nested cross-validation approach combining a feature selection technique with three predictive models: random forest, support vector machine (SVM) with Gaussian Kernel, and SVM with polynomial kernel. The polynomial SVM yielded the best classification performance. Receiver operating characteristic curves provided area-under-the-curve (AUC) (mean±standard deviation) of 0.86±0.06 on LGE MRI using 72 features; AMI sensitivity=0.81±0.08 and specificity=0.84±0.09. On cine MRI, AUC=0.82±0.06 using 75 features; AMI sensitivity=0.79±0.10 and specificity=0.80±0.10. We concluded that texture analysis can be used for differentiation of AMI from CMI on cardiac LGE MRI, and also on standard cine sequences in which the infarction is visually imperceptible in most cases. Copyright © 2017 Elsevier B.V. All rights reserved.

  12. Functional measurements based on feature tracking of cine magnetic resonance images identify left ventricular segments with myocardial scar

    Directory of Open Access Journals (Sweden)

    Nylander Eva

    2009-11-01

    Full Text Available Abstract Background The aim of the study was to perform a feature tracking analysis on cine magnetic resonance (MR images to elucidate if functional measurements of the motion of the left ventricular wall may detect scar defined with gadolinium enhanced MR. Myocardial contraction can be measured in terms of the velocity, displacement and local deformation (strain of a particular myocardial segment. Contraction of the myocardial wall will be reduced in the presence of scar and as a consequence of reduced myocardial blood flow. Methods Thirty patients (3 women and 27 men were selected based on the presence or absence of extensive scar in the anteroseptal area of the left ventricle. The patients were investigated in stable clinical condition, 4-8 weeks post ST-elevation myocardial infarction treated with percutaneous coronary intervention. Seventeen had a scar area >75% in at least one anteroseptal segment (scar and thirteen had scar area Results In the scar patients, segments with scar showed lower functional measurements than remote segments. Radial measurements of velocity, displacement and strain performed better in terms of receiver-operator-characteristic curves (ROC than the corresponding longitudinal measurements. The best area-under-curve was for radial strain, 0.89, where a cut-off value of 38.8% had 80% sensitivity and 86% specificity for the detection of a segment with scar area >50%. As a percentage of the mean, intraobserver variability was 16-14-26% for radial measurements of displacement-velocity-strain and corresponding interobserver variability was 13-12-18%. Conclusion Feature tracking analysis of cine-MR displays velocity, displacement and strain in the radial and longitudinal direction and may be used for the detection of transmural scar. The accuracy and repeatability of the radial functional measurements is satisfactory and global measures agree.

  13. Cardiac cine MRI: Comparison of 1.5 T, non-enhanced 3.0 T and blood pool enhanced 3.0 T imaging

    International Nuclear Information System (INIS)

    Gerretsen, S.C.; Versluis, B.; Bekkers, S.C.A.M.; Leiner, T.

    2008-01-01

    Introduction: Cardiac cine imaging using balanced steady state free precession sequences (bSSFP) suffers from artefacts at 3.0 T. We compared bSSFP cardiac cine imaging at 1.5 T with gradient echo imaging at 3.0 T with and without a blood pool contrast agent. Materials and methods: Eleven patients referred for cardiac cine imaging underwent imaging at 1.5 T and 3.0 T. At 3.0 T images were acquired before and after administration of 0.03 mmol/kg gadofosveset. Blood pool signal-to-noise ratio (SNR), temporal variations in SNR, ejection fraction and myocardial mass were compared. Subjective image quality was scored on a four-point scale. Results: Blood pool SNR increased with more than 75% at 3.0 T compared to 1.5 T (p < 0.001); after contrast administration at 3.0 T SNR increased with 139% (p < 0.001). However, variations in blood pool SNR at 3.0 T were nearly three times as high versus those at 1.5 T in the absence of contrast medium (p < 0.001); after contrast administration this was reduced to approximately a factor 1.4 (p = 0.21). Saturation artefacts led to significant overestimation of ejection fraction in the absence of contrast administration (1.5 T: 44.7 ± 3.1 vs. 3.0 T: 50.7 ± 4.2 [p = 0.04] vs. 3.0 T post contrast: 43.4 ± 2.9 [p = 0.55]). Subjective image quality was highest for 1.5 T (2.8 ± 0.3), and lowest for non-enhanced 3.0 T (1.7 ± 0.6; p = 0.006). Conclusions: GRE cardiac cine imaging at 3.0 T after injection of the blood pool agent gadofosveset leads to improved objective and subjective cardiac cine image quality at 3.0 T and to the same conclusions regarding cardiac ejection fraction compared to bSSFP imaging at 1.5 T

  14. SU-F-T-240: EPID-Based Quality Assurance for Dosimetric Credentialing

    Energy Technology Data Exchange (ETDEWEB)

    Miri, N [University of Newcastle, Newcastle, NSW (Australia); Lehmann, J [Calvary Mater Newcastle, Newcastle, NSW (Australia); Vial, P [Liverpool Hospital, Sydney, NSW (Australia); Greer, P [Calvary Mater Newcastle, Newcastle, NSW (Australia); University of Newcastle, Newcastle, NSW (Australia)

    2016-06-15

    Purpose: We propose a novel dosimetric audit method for clinical trials using EPID measurements at each center and a standardized EPID to dose conversion algorithm. The aim of this work is to investigate the applicability of the EPID method to different linear accelerator, EPID and treatment planning system (TPS) combinations. Methods: Combination of delivery and planning systems were three Varian linacs including one Pinnacle and two Eclipse TPS and, two ELEKTA linacs including one Pinnacle and one Monaco TPS. All Varian linacs had the same EPID structure and similarly for the ELEKTA linacs. Initially, dose response of the EPIDs was investigated by acquiring integrated pixel value (IPV) of the central area of 10 cm2 images versus MUs, 5-400 MU. Then, the EPID to dose conversion was investigated for different system combinations. Square field size images, 2, 3, 4, 6, 10, 15, 20, 25 cm2 acquired by all systems were converted to dose at isocenter of a virtual flat phantom then the dose was compared to the corresponding TPS dose. Results: All EPIDs showed a relatively linear behavior versus MU except at low MUs which showed irregularities probably due to initial inaccuracies of irradiation. Furthermore, for all the EPID models, the model predicted TPS dose with a mean dose difference percentage of 1.3. However the model showed a few inaccuracies for ELEKTA EPID images at field sizes larger than 20 cm2. Conclusion: The EPIDs demonstrated similar behavior versus MU and the model was relatively accurate for all the systems. Therefore, the model could be employed as a global dosimetric method to audit clinical trials. Funding has been provided from Department of Radiation Oncology, TROG Cancer Research and the University of Newcastle. Narges Miri is a recipient of a University of Newcastle postgraduate scholarship.

  15. Towards radiological diagnosis of abdominal adhesions based on motion signatures derived from sequences of cine-MRI images.

    Science.gov (United States)

    Fenner, John; Wright, Benjamin; Emberey, Jonathan; Spencer, Paul; Gillott, Richard; Summers, Angela; Hutchinson, Charles; Lawford, Pat; Brenchley, Paul; Bardhan, Karna Dev

    2014-06-01

    This paper reports novel development and preliminary application of an image registration technique for diagnosis of abdominal adhesions imaged with cine-MRI (cMRI). Adhesions can severely compromise the movement and physiological function of the abdominal contents, and their presence is difficult to detect. The image registration approach presented here is designed to expose anomalies in movement of the abdominal organs, providing a movement signature that is indicative of underlying structural abnormalities. Validation of the technique was performed using structurally based in vitro and in silico models, supported with Receiver Operating Characteristic (ROC) methods. For the more challenging cases presented to the small cohort of 4 observers, the AUC (area under curve) improved from a mean value of 0.67 ± 0.02 (without image registration assistance) to a value of 0.87 ± 0.02 when image registration support was included. Also, in these cases, a reduction in time to diagnosis was observed, decreasing by between 20% and 50%. These results provided sufficient confidence to apply the image registration diagnostic protocol to sample magnetic resonance imaging data from healthy volunteers as well as a patient suffering from encapsulating peritoneal sclerosis (an extreme form of adhesions) where immobilization of the gut by cocooning of the small bowel is observed. The results as a whole support the hypothesis that movement analysis using image registration offers a possible method for detecting underlying structural anomalies and encourages further investigation. Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  16. Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithm

    International Nuclear Information System (INIS)

    Wintersperger, Bernd J.; Nikolaou, Konstantin; Dietrich, Olaf; Reiser, Maximilian F.; Schoenberg, Stefan O.; Rieber, Johannes; Nittka, Matthias

    2003-01-01

    The purpose of this study was to test parallel imaging techniques for improvement of temporal resolution in multislice single breath-hold real-time cine steady-state free precession (SSFP) in comparison with standard segmented single-slice SSFP techniques. Eighteen subjects were examined on a 1.5-T scanner using a multislice real-time cine SSFP technique using the GRAPPA algorithm. Global left ventricular parameters (EDV, ESV, SV, EF) were evaluated and results compared with a standard segmented single-slice SSFP technique. Results for EDV (r=0.93), ESV (r=0.99), SV (r=0.83), and EF (r=0.99) of real-time multislice SSFP imaging showed a high correlation with results of segmented SSFP acquisitions. Systematic differences between both techniques were statistically non-significant. Single breath-hold multislice techniques using GRAPPA allow for improvement of temporal resolution and for accurate assessment of global left ventricular functional parameters. (orig.)

  17. Influence of high magnetic field strengths and parallel acquisition strategies on image quality in cardiac 2D CINE magnetic resonance imaging: comparison of 1.5 T vs. 3.0 T

    International Nuclear Information System (INIS)

    Gutberlet, Matthias; Schwinge, Kerstin; Freyhardt, Patrick; Spors, Birgit; Grothoff, Matthias; Denecke, Timm; Luedemann, Lutz; Felix, Roland; Noeske, Ralph; Niendorf, Thoralf

    2005-01-01

    The aim of this paper is to examine signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and image quality of cardiac CINE imaging at 1.5 T and 3.0 T. Twenty volunteers underwent cardiac magnetic resonance imaging (MRI) examinations using a 1.5-T and a 3.0-T scanner. Three different sets of breath-held, electrocardiogram-gated (ECG) CINE imaging techniques were employed, including: (1) unaccelerated SSFP (steady state free precession), (2) accelerated SSFP imaging and (3) gradient-echo-based myocardial tagging. Two-dimensional CINE SSFP at 3.0 T revealed an SNR improvement of 103% and a CNR increase of 19% as compared to the results obtained at 1.5 T. The SNR reduction in accelerated 2D CINE SSFP imaging was larger at 1.5 T (37%) compared to 3.0 T (26%). The mean SNR and CNR increase at 3.0 T obtained for the tagging sequence was 88% and 187%, respectively. At 3.0 T, the duration of the saturation bands persisted throughout the entire cardiac cycle. For comparison, the saturation bands were significantly diminished at 1.5 T during end-diastole. For 2D CINE SSFP imaging, no significant difference in the left ventricular volumetry and in the overall image quality was obtained. For myocardial tagging, image quality was significantly improved at 3.0 T. The SNR reduction in accelerated SSFP imaging was overcompensated by the increase in the baseline SNR at 3.0 T and did not result in any image quality degradation. For cardiac tagging techniques, 3.0 T was highly beneficial, which holds the promise to improve its diagnostic value. (orig.)

  18. Two-dimensional XD-GRASP provides better image quality than conventional 2D cardiac cine MRI for patients who cannot suspend respiration.

    Science.gov (United States)

    Piekarski, Eve; Chitiboi, Teodora; Ramb, Rebecca; Latson, Larry A; Bhatla, Puneet; Feng, Li; Axel, Leon

    2018-02-01

    Residual respiratory motion degrades image quality in conventional cardiac cine MRI (CCMRI). We evaluated whether a free-breathing (FB) radial imaging CCMRI sequence with compressed sensing reconstruction [extradimensional (e.g. cardiac and respiratory phases) golden-angle radial sparse parallel, or XD-GRASP] could provide better image quality than a conventional Cartesian breath-held (BH) sequence in an unselected population of patients undergoing clinical CCMRI. One hundred one patients who underwent BH and FB imaging in a midventricular short-axis plane at a matching location were included. Visual and quantitative image analysis was performed by two blinded experienced readers, using a five-point qualitative scale to score overall image quality and visual signal-to-noise ratio (SNR) grade, with measures of noise and sharpness. End-diastolic and end-systolic left ventricular areas were also measured and compared for both BH and FB images. Image quality was generally better with the BH cines (overall quality grade for BH vs FB images 4 vs 2.9, p XD-GRASP CCMRI was visually inferior to conventional BH CCMRI in general, it provided improved image quality in the subgroup of patients with respiratory-motion-induced artifacts on BH images.

  19. SU-F-303-11: Implementation and Applications of Rapid, SIFT-Based Cine MR Image Binning and Region Tracking

    Energy Technology Data Exchange (ETDEWEB)

    Mazur, T; Wang, Y; Fischer-Valuck, B; Acharya, S; Kashani, R; Li, H; Yang, D; Zoberi, I; Thomas, M; Mutic, S; Li, H [Washington University School of Medicine, St. Louis, MO (United States)

    2015-06-15

    Purpose: To develop a novel and rapid, SIFT-based algorithm for assessing feature motion on cine MR images acquired during MRI-guided radiotherapy treatments. In particular, we apply SIFT descriptors toward both partitioning cine images into respiratory states and tracking regions across frames. Methods: Among a training set of images acquired during a fraction, we densely assign SIFT descriptors to pixels within the images. We cluster these descriptors across all frames in order to produce a dictionary of trackable features. Associating the best-matching descriptors at every frame among the training images to these features, we construct motion traces for the features. We use these traces to define respiratory bins for sorting images in order to facilitate robust pixel-by-pixel tracking. Instead of applying conventional methods for identifying pixel correspondences across frames we utilize a recently-developed algorithm that derives correspondences via a matching objective for SIFT descriptors. Results: We apply these methods to a collection of lung, abdominal, and breast patients. We evaluate the procedure for respiratory binning using target sites exhibiting high-amplitude motion among 20 lung and abdominal patients. In particular, we investigate whether these methods yield minimal variation between images within a bin by perturbing the resulting image distributions among bins. Moreover, we compare the motion between averaged images across respiratory states to 4DCT data for these patients. We evaluate the algorithm for obtaining pixel correspondences between frames by tracking contours among a set of breast patients. As an initial case, we track easily-identifiable edges of lumpectomy cavities that show minimal motion over treatment. Conclusions: These SIFT-based methods reliably extract motion information from cine MR images acquired during patient treatments. While we performed our analysis retrospectively, the algorithm lends itself to prospective motion

  20. Reproducibility of small animal cine and scar cardiac magnetic resonance imaging using a clinical 3.0 tesla system

    International Nuclear Information System (INIS)

    Manka, Robert; Jahnke, Cosima; Hucko, Thomas; Dietrich, Thore; Gebker, Rolf; Schnackenburg, Bernhard; Graf, Kristof; Paetsch, Ingo

    2013-01-01

    To evaluate the inter-study, inter-reader and intra-reader reproducibility of cardiac cine and scar imaging in rats using a clinical 3.0 Tesla magnetic resonance (MR) system. Thirty-three adult rats (Sprague–Dawley) were imaged 24 hours after surgical occlusion of the left anterior descending coronary artery using a 3.0 Tesla clinical MR scanner (Philips Healthcare, Best, The Netherlands) equipped with a dedicated 70 mm solenoid receive-only coil. Left-ventricular (LV) volumes, mass, ejection fraction and amount of myocardial scar tissue were measured. Intra-and inter-observer reproducibility was assessed in all animals. In addition, repeat MR exams were performed in 6 randomly chosen rats within 24 hours to assess inter-study reproducibility. The MR imaging protocol was successfully completed in 32 (97%) animals. Bland-Altman analysis demonstrated high intra-reader reproducibility (mean bias%: LV end-diastolic volume (LVEDV), -1.7%; LV end-systolic volume (LVESV), -2.2%; LV ejection fraction (LVEF), 1.0%; LV mass, -2.7%; and scar mass, -1.2%) and high inter-reader reproducibility (mean bias%: LVEDV, 3.3%; LVESV, 6.2%; LVEF, -4.8%; LV mass, -1.9%; and scar mass, -1.8%). In addition, a high inter-study reproducibility was found (mean bias%: LVEDV, 0.1%; LVESV, -1.8%; LVEF, 1.0%; LV mass, -4.6%; and scar mass, -6.2%). Cardiac MR imaging of rats yielded highly reproducible measurements of cardiac volumes/function and myocardial infarct size on a clinical 3.0 Tesla MR scanner system. Consequently, more widely available high field clinical MR scanners can be employed for small animal imaging of the heart e.g. when aiming at serial assessments during therapeutic intervention studies

  1. Free-breathing black-blood CINE fast-spin echo imaging for measuring abdominal aortic wall distensibility: a feasibility study

    Science.gov (United States)

    Lin, Jyh-Miin; Patterson, Andrew J.; Chao, Tzu-Cheng; Zhu, Chengcheng; Chang, Hing-Chiu; Mendes, Jason; Chung, Hsiao-Wen; Gillard, Jonathan H.; Graves, Martin J.

    2017-05-01

    The paper reports a free-breathing black-blood CINE fast-spin echo (FSE) technique for measuring abdominal aortic wall motion. The free-breathing CINE FSE includes the following MR techniques: (1) variable-density sampling with fast iterative reconstruction; (2) inner-volume imaging; and (3) a blood-suppression preparation pulse. The proposed technique was evaluated in eight healthy subjects. The inner-volume imaging significantly reduced the intraluminal artifacts of respiratory motion (p  =  0.015). The quantitative measurements were a diameter of 16.3  ±  2.8 mm and wall distensibility of 2.0  ±  0.4 mm (12.5  ±  3.4%) and 0.7  ±  0.3 mm (4.1  ±  1.0%) for the anterior and posterior walls, respectively. The cyclic cross-sectional distensibility was 35  ±  15% greater in the systolic phase than in the diastolic phase. In conclusion, we developed a feasible CINE FSE method to measure the motion of the abdominal aortic wall, which will enable clinical scientists to study the elasticity of the abdominal aorta.

  2. Detection of diminished response to cold pressor test in smokers: assessment using phase-contrast cine magnetic resonance imaging of the coronary sinus.

    Science.gov (United States)

    Kato, Shingo; Kitagawa, Kakuya; Yoon, Yeonyee E; Nakajima, Hiroshi; Nagata, Motonori; Takase, Shinichi; Nakamori, Shiro; Ito, Masaaki; Sakuma, Hajime

    2014-04-01

    The purposes of this study were to evaluate the reproducibility for measuring the cold pressor test (CPT)-induced myocardial blood flow (MBF) alteration using phase-contrast (PC) cine MRI, and to determine if this approach could detect altered MBF response to CPT in smokers. After obtaining informed consent, ten healthy male non-smokers (mean age: 28±5 years) and ten age-matched male smokers (smoking duration ≥5 years, mean age: 28±3 years) were examined in this institutional review board approved study. Breath-hold PC cine MR images of the coronary sinus were obtained with a 3T MR imager with 32 channel coils at rest and during a CPT performed after immersing one foot in ice water. MBF was calculated as coronary sinus flow divided by the left ventricular (LV) mass which was given as a total LV myocardial volume measured on cine MRI multiplied by the specific gravity (1.05 g/mL). In non-smokers, MBF was 0.86±0.25 mL/min/g at rest, with a significant increase to 1.20±0.36 mL/min/g seen during CPT (percentage change of MBF (∆MBF (%)); 39.2%±14.4%, psmokers and non-smokers for resting MBF (0.85±0.32 mL/min/g, p=0.91). However, ∆MBF (%) in smokers was significantly reduced (-4.0±32.2% vs. 39.2±14.4%, p=0.011). PC cine MRI can be used to reproducibly quantify MBF response to CPT and to detect impaired flow response in smokers. This MR approach may be useful for monitoring the sequential change of coronary blood flow in various potentially pathologic conditions and for investigating its relationship with cardiovascular risk. Copyright © 2014 Elsevier Inc. All rights reserved.

  3. Using fluence separation to account for energy spectra dependence in computing dosimetric a-Si EPID images for IMRT fields

    International Nuclear Information System (INIS)

    Li Weidong; Siebers, Jeffrey V.; Moore, Joseph A.

    2006-01-01

    This study develops a method to improve the dosimetric accuracy of computed images for an amorphous silicon flat-panel imager. Radially dependent kernels derived from Monte Carlo simulations are convolved with the treatment-planning system's energy fluence. Multileaf collimator (MLC) beam hardening is accounted for by having separate kernels for open and blocked portions of MLC fields. Field-size-dependent output factors are used to account for the field-size dependence of scatter within the imager. Gamma analysis was used to evaluate open and sliding window test fields and intensity modulated patient fields. For each tested field, at least 99.6% of the points had γ<1 with a 3%, 3-mm criteria. With a 2%, 2-mm criteria, between 81% and 100% of points had γ<1. Patient intensity modulated test fields had 94%-100% of the points with γ<1 with a 2%, 2-mm criteria for all six fields tested. This study demonstrates that including the dependencies of kernel and fluence on radius and beam hardening in the convolution improves its accuracy compared with the use of radial and beam-hardening independent kernels; it also demonstrates that the resultant accuracy of the convolution method is sufficient for pretreatment, intensity modulated patient field verification

  4. An EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments

    International Nuclear Information System (INIS)

    Rowshanfarzad, Pejman; McGarry, Conor K; Barnes, Michael P; Sabet, Mahsheed; Ebert, Martin A

    2014-01-01

    In modern radiotherapy, it is crucial to monitor the performance of all linac components including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method has been introduced in conjunction with an algorithm to investigate the stability of these systems during arc treatments with the aim of ensuring the accuracy of linac mechanical performance. The Varian EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in MLC carriages as a result of linac rotation were separately investigated by acquisition of EPID images of a simple phantom comprised of 5 ball-bearings during arc delivery. A fast and robust software package was developed for automated analysis of image data. Twelve Varian linacs of different models were investigated. The average EPID sag was within 1 mm for all tested linacs. All machines showed less than 1 mm gantry sag. Changes in SDD values were within 1.7 mm except for three linacs of one centre which were within 9 mm. Values of EPID skewness and tilt were negligible in all tested linacs. The maximum sag in MLC leaf bank assemblies was around 1 mm. The EPID sag showed a considerable improvement in TrueBeam linacs. The methodology and software developed in this study provide a simple tool for effective investigation of the behaviour of linac components with gantry rotation. It is reproducible and accurate and can be easily performed as a routine test in clinics

  5. In-vivo quantification of wall motion in cerebral aneurysms from 2D cine phase contrast magnetic resonance images

    Energy Technology Data Exchange (ETDEWEB)

    Karmonik, C. [The Methodist Hospital Research Inst., Houston (United States); Diaz, O.; Klucznik, R. [The Methodist Hospital, Houston (United States); Grossman, R. [The Methodist Hospital, Houston (United States). Neurosurgery

    2010-02-15

    Purpose: The quantification of wall motion in cerebral aneurysms is of interest for the assessment of aneurysmal rupture risk, for providing boundary conditions for computational simulations and as a validation tool for theoretical models. Materials and Methods: 2D cine phase contrast magnetic resonance imaging (2D pcMRI) in combination with quantitative magnetic resonance angiography (QMRA) was evaluated for measuring wall motion in 7 intracranial aneurysms. In each aneurysm, 2 (in one case 3) cross sections, oriented approximately perpendicular to each other, were measured. Results: The maximum aneurysmal wall distention ranged from 0.16 mm to 1.6 mm (mean 0.67 mm), the maximum aneurysmal wall contraction was -1.91 mm to -0.34 mm (mean 0.94 mm), and the average wall displacement ranged from 0.04 mm to 0.31 mm (mean 0.15 mm). Statistically significant correlations between average wall displacement and the shape of inflow curves (p-value < 0.05) were found in 7 of 15 cross sections; statistically significant correlations between the displacement of the luminal boundary center point and the shape of inflow curves (p-value < 0.05) were found in 6 of 15 cross sections. Conclusion: 2D pcMRI in combination with QMRA is capable of visualizing and quantifying wall motion in cerebral aneurysms. However, application of this technique is currently restricted by its limited spatial resolution. (orig.)

  6. An improved Monte-Carlo model of the Varian EPID separating support arm and rear-housing backscatter

    International Nuclear Information System (INIS)

    Monville, M E; Greer, P B; Kuncic, Z

    2014-01-01

    Previous investigators of EPID dosimetric properties have ascribed the backscatter, that contaminates dosimetric EPID images, to its supporting arm. Accordingly, Monte-Carlo (MC) EPID models have approximated the backscatter signal from the layers under the detector and the robotic support arm using either uniform or non-uniform solid water slabs, or through convolutions with back-scatter kernels. The aim of this work is to improve the existent MC models by measuring and modelling the separate backscatter contributions of the robotic arm and the rear plastic housing of the EPID. The EPID plastic housing is non-uniform with a 11.9 cm wide indented section that runs across the cross-plane direction in the superior half of the EPID which is 1.75 cm closer to the EPID sensitive layer than the rest of the housing. The thickness of the plastic housing is 0.5 cm. Experiments were performed with and without the housing present by removing all components of the EPID from the housing. The robotic support arm was not present for these measurements. A MC model of the linear accelerator and the EPID was modified to include the rear-housing indentation and results compared to the measurement. The rear housing was found to contribute a maximum of 3% additional signal. The rear housing contribution to the image is non-uniform in the in-plane direction with 2% asymmetry across the central 20 cm of an image irradiating the entire detector. The MC model was able to reproduce this non-uniform contribution. The EPID rear housing contributes a non-uniform backscatter component to the EPID image, which has not been previously characterized. This has been incorporated into an improved MC model of the EPID.

  7. Segmentation of left ventricle myocardium in porcine cardiac cine MR images using a hybrid of fully convolutional neural networks and convolutional LSTM

    Science.gov (United States)

    Zhang, Dongqing; Icke, Ilknur; Dogdas, Belma; Parimal, Sarayu; Sampath, Smita; Forbes, Joseph; Bagchi, Ansuman; Chin, Chih-Liang; Chen, Antong

    2018-03-01

    In the development of treatments for cardiovascular diseases, short axis cardiac cine MRI is important for the assessment of various structural and functional properties of the heart. In short axis cardiac cine MRI, Cardiac properties including the ventricle dimensions, stroke volume, and ejection fraction can be extracted based on accurate segmentation of the left ventricle (LV) myocardium. One of the most advanced segmentation methods is based on fully convolutional neural networks (FCN) and can be successfully used to do segmentation in cardiac cine MRI slices. However, the temporal dependency between slices acquired at neighboring time points is not used. Here, based on our previously proposed FCN structure, we proposed a new algorithm to segment LV myocardium in porcine short axis cardiac cine MRI by incorporating convolutional long short-term memory (Conv-LSTM) to leverage the temporal dependency. In this approach, instead of processing each slice independently in a conventional CNN-based approach, the Conv-LSTM architecture captures the dynamics of cardiac motion over time. In a leave-one-out experiment on 8 porcine specimens (3,600 slices), the proposed approach was shown to be promising by achieving average mean Dice similarity coefficient (DSC) of 0.84, Hausdorff distance (HD) of 6.35 mm, and average perpendicular distance (APD) of 1.09 mm when compared with manual segmentations, which improved the performance of our previous FCN-based approach (average mean DSC=0.84, HD=6.78 mm, and APD=1.11 mm). Qualitatively, our model showed robustness against low image quality and complications in the surrounding anatomy due to its ability to capture the dynamics of cardiac motion.

  8. On the system of cine-angiography

    International Nuclear Information System (INIS)

    Wakamatsu, Koji; Togi, Hideaki; Yokoyama, Hironori

    1979-01-01

    National Institute of Circulatory Disease Centre have four angiography apparatuses which deal with more than 80 cases of cerebral blood vessel and 100 cases of heart blood vessel each month. Most of the angiography is cine-photography. Five more angiography apparatuses are expected to be operated in operation and RI examination rooms in the future. The problems on cine-angiography system were discussed as follows: (1) The proper dose must be chosen in order to obtain good quality of cine-images. (2) The cine resolution depends much on dose. (3) The lower X-ray tube voltage can give better contrast. (4) Suitable capacity for an X-ray tube is over 60 kW at 2 sec rating. (5) Cine-autophotography requires rapid response and a circuit for lock. (6) Influence of side scattering can be solved by a blanking circuit in biplane cine-photography. (7) Self-developing is desirable to improve the quality of cine-images. (Kobatake, H.)

  9. Cine-Club

    CERN Multimedia

    Cine-Club

    2013-01-01

    The CERN CineClub is a meeting point every Thursday for anyone who wants to discover, discuss, and appreciate cinema with people who are truly interested in the art of the moving image. In the actual context where downloading became easier than cooking an egg, one might wonder why bother when it can be watched on any computer. Well, cinema is not an individual activity. What the CineClub offers is a big screen with good quality image and sound. It also offers a careful selection, for allowing people to (re-)discover new and forgotten authors. Finally, it offers the opportunity for meeting people sharing a common interest. So come every Thursday, and see by yourself that watching films can actually be a real party! Every Thursday at 20 h 00 Council Chamber, 503/1-001 Programme June 2013 Science Fiction Cycle “Imagination is more important than knowledge. For knowledge is limited, whereas imagination embraces the entire world, stimulating progress, giving birth to evolution.” 6th June 2...

  10. CINE CLUB

    CERN Multimedia

    Ciné Club

    2009-01-01

    Main Auditorium CERN jeudi 17 décembre 2009 à 18 h 15 à l’Amphithéâtre principal du CERN (bâtiment 500)   Comme chaque année avant Noël, le CineClub du CERN est heureux d’inviter petits et grands à une projection gratuite du film   LE MONDE DE NEMO (FINDING NEMO) (USA, 2003, Andrew Stanton and Lee Unkrich)   Dans les eaux tropicales de la Grande Barrière de corail, un poisson-clown du nom de Marin mène une existence paisible avec son fils unique, Nemo. Redoutant l’océan et ses risques imprévisibles, il fait de son mieux pour protéger son fils. Comme tous les petits poissons de son âge, celui-ci rêve pourtant d’explorer les mystérieux récifs. Lorsque Nemo disparaît, Marin devient malgré lui le héros d’une qu&a...

  11. Cine club

    CERN Multimedia

    Ciné club

    2010-01-01

    Monday 13 December 2010 at 18:30 / Lundi 13 Décembre 2010 à 18:30 CERN Main Auditorium / Amphithéâtre Principal Comme chaque année avant Noël, le CINE-CLUB du CERN est heureux d’inviter petits et grands à une projection gratuite du film.   Ponyo sur la falaise (Japon, 2008, Hayao Miyazaki) Le petit Sosuke vit avec sa mère sur une haute falaise surplombant la mer. Un beau jour, il découvre sur la plage caillouteuse Ponyo, une petite fille poisson. Ponyo est si fascinée par Sosuke et le monde terrestre que son désir le plus cher est de devenir un être humain. Mais Fujimoto, son magicien de père, n’est pas du tout d’accord avec cette idée et il la force à regagner les profondeurs de l’océan. Bien décidée à revoir Sosuke, Ponyo s’&...

  12. Flow velocity and volume measurement of superior and inferior mesenteric artery with cine phase contrast magnetic resonance imaging

    International Nuclear Information System (INIS)

    Naganawa, Shinji; Cooper, T.G.; Jenner, G.; Potchen, E.J.; Ishigaki, Takeo.

    1994-01-01

    The flow velocity and volume of the superior and inferior mesenteric arteries (SMA, IMA) were measured with cine phase contrast magnetic resonance (MR) imaging in five healthy volunteers. Each volunteer was first measured in a fasting state, and then one, two, and three hours after a meal. The average SMA flow volume of the volunteers was 230.3±46.8 ml/min (mean±standard error) during the fasting state, and 714.7±207.7 ml/min, 339.2±85.7 ml/min, and 263.8±21.0 ml/min, respectively, at one, two, and three hours postmeal. The increase at one hour postmeal was statistically significant (p<0.05). The corresponding flow measurements in the IMA were 63.1±11.2 ml/min, 67.6±11.2 ml/min, 57.9±8.6 ml/min, and 53.2±6.8 ml/min. These values do not represent a statistically significant flow volume change in the IMA. In all volunteers, the SMA volumetric flow increased the most one hour after the food challenge (72-400% relative to baseline). Diastolic velocity in the SMA increased significantly one hour postmeal, but systolic velocity did not change significantly. The IMA did not demonstrate a significant change in either systolic or diastolic velocity. The difference between the SMA and IMA in the way of reacting against the food challenge is thought to represent the difference between the requirements of small and large intestine for blood supply after the food challenge. These data demonstrate the possibility of this modality for the assessment of conditions such as chronic mesenteric ischemia. (author)

  13. How to perform and interpret cine MR enterography.

    Science.gov (United States)

    Wnorowski, Amelia M; Guglielmo, Flavius F; Mitchell, Donald G

    2015-11-01

    Magnetic resonance (MR) enterography has become a fundamental tool for small bowel evaluation. Multiphasic cine imaging is a useful component of MR enterography evaluation because it provides functional information about bowel motility. Cine MR enterography can be used to evaluate for strictures and adhesions. Bowel motility evaluation has been shown to increase pathologic lesion detection in Crohn's disease and has been incorporated into disease activity scoring systems. Currently, cine MR enterography remains underutilized. The purpose of this article is to outline how to perform and interpret cine MR enterography. The authors describe how to perform a multiphasic balanced steady state free precession sequence using different MR systems and give practical advice on how to display and interpret the cine sequence. Sample cases illustrate how the cine sequence complements standard MR enterography evaluation with T2 -weighted, contrast-enhanced T1 -weighted, and diffusion-weighted imaging. © 2015 Wiley Periodicals, Inc.

  14. SU-F-J-114: On-Treatment Imagereconstruction Using Transit Images of Treatment Beams Through Patient and Thosethrough Planning CT Images

    Energy Technology Data Exchange (ETDEWEB)

    Lee, H; Cho, S [KAIST, Yuseong-gu, Daejeon (Korea, Republic of); Cheong, K [Hallym University Sacred Heart Hospital, Anyang (Korea, Republic of); Jung, J [East Carolina University Greenville, NC (United States); Jung, S [Samsung Medical Cener, Gangnam-gu, Seoul (Korea, Republic of); Kim, J [Yonsei Cancer Center, Seoul (Korea, Republic of); Yeo, I [Loma Linda University Medical Center, Loma Linda, CA (United States)

    2016-06-15

    Purpose: To reconstruct patient images at the time of radiation delivery using measured transit images of treatment beams through patient and calculated transit images through planning CT images. Methods: We hypothesize that the ratio of the measured transit images to the calculated images may provide changed amounts of the patient image between times of planning CT and treatment. To test, we have devised lung phantoms with a tumor object (3-cm diameter) placed at iso-center (simulating planning CT) and off-center by 1 cm (simulating treatment). CT images of the two phantoms were acquired; the image of the off-centered phantom, unavailable clinically, represents the reference on-treatment image in the image quality of planning CT. Cine-transit images through the two phantoms were also acquired in EPID from a non-modulated 6 MV beam when the gantry was rotated 360 degrees; the image through the centered phantom simulates calculated image. While the current study is a feasibility study, in reality our computational EPID model can be applicable in providing accurate transit image from MC simulation. Changed MV HU values were reconstructed from the ratio between two EPID projection data, converted to KV HU values, and added to the planning CT, thereby reconstructing the on-treatment image of the patient limited to the irradiated region of the phantom. Results: The reconstructed image was compared with the reference image. Except for local HU differences>200 as a maximum, excellent agreement was found. The average difference across the entire image was 16.2 HU. Conclusion: We have demonstrated the feasibility of a method of reconstructing on-treatment images of a patient using EPID image and planning CT images. Further studies will include resolving the local HU differences and investigation on the dosimetry impact of the reconstructed image.

  15. SU-F-J-114: On-Treatment Imagereconstruction Using Transit Images of Treatment Beams Through Patient and Thosethrough Planning CT Images

    International Nuclear Information System (INIS)

    Lee, H; Cho, S; Cheong, K; Jung, J; Jung, S; Kim, J; Yeo, I

    2016-01-01

    Purpose: To reconstruct patient images at the time of radiation delivery using measured transit images of treatment beams through patient and calculated transit images through planning CT images. Methods: We hypothesize that the ratio of the measured transit images to the calculated images may provide changed amounts of the patient image between times of planning CT and treatment. To test, we have devised lung phantoms with a tumor object (3-cm diameter) placed at iso-center (simulating planning CT) and off-center by 1 cm (simulating treatment). CT images of the two phantoms were acquired; the image of the off-centered phantom, unavailable clinically, represents the reference on-treatment image in the image quality of planning CT. Cine-transit images through the two phantoms were also acquired in EPID from a non-modulated 6 MV beam when the gantry was rotated 360 degrees; the image through the centered phantom simulates calculated image. While the current study is a feasibility study, in reality our computational EPID model can be applicable in providing accurate transit image from MC simulation. Changed MV HU values were reconstructed from the ratio between two EPID projection data, converted to KV HU values, and added to the planning CT, thereby reconstructing the on-treatment image of the patient limited to the irradiated region of the phantom. Results: The reconstructed image was compared with the reference image. Except for local HU differences>200 as a maximum, excellent agreement was found. The average difference across the entire image was 16.2 HU. Conclusion: We have demonstrated the feasibility of a method of reconstructing on-treatment images of a patient using EPID image and planning CT images. Further studies will include resolving the local HU differences and investigation on the dosimetry impact of the reconstructed image.

  16. EPID-based verification of the MLC performance for dynamic IMRT and VMAT

    International Nuclear Information System (INIS)

    Rowshanfarzad, Pejman; Sabet, Mahsheed; Barnes, Michael P.; O’Connor, Daryl J.; Greer, Peter B.

    2012-01-01

    Purpose: In advanced radiotherapy treatments such as intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT), verification of the performance of the multileaf collimator (MLC) is an essential part of the linac QA program. The purpose of this study is to use the existing measurement methods for geometric QA of the MLCs and extend them to more comprehensive evaluation techniques, and to develop dedicated robust algorithms to quantitatively investigate the MLC performance in a fast, accurate, and efficient manner. Methods: The behavior of leaves was investigated in the step-and-shoot mode by the analysis of integrated electronic portal imaging device (EPID) images acquired during picket fence tests at fixed gantry angles and arc delivery. The MLC was also studied in dynamic mode by the analysis of cine EPID images of a sliding gap pattern delivered in a variety of conditions including different leaf speeds, deliveries at fixed gantry angles or in arc mode, and changing the direction of leaf motion. The accuracy of the method was tested by detection of the intentionally inserted errors in the delivery patterns. Results: The algorithm developed for the picket fence analysis was able to find each individual leaf position, gap width, and leaf bank skewness in addition to the deviations from expected leaf positions with respect to the beam central axis with sub-pixel accuracy. For the three tested linacs over a period of 5 months, the maximum change in the gap width was 0.5 mm, the maximum deviation from the expected leaf positions was 0.1 mm and the MLC skewness was up to 0.2°. The algorithm developed for the sliding gap analysis could determine the velocity and acceleration/deceleration of each individual leaf as well as the gap width. There was a slight decrease in the accuracy of leaf performance with increasing leaf speeds. The analysis results were presented through several graphs. The accuracy of the method was assessed as 0.01 mm

  17. Accurate estimation of global and regional cardiac function by retrospectively gated multidetector row computed tomography. Comparison with cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Belge, Benedicte; Pasquet, Agnes; Vanoverschelde, Jean-Louis J.; Coche, Emmanuel; Gerber, Bernhard L.

    2006-01-01

    Retrospective reconstruction of ECG-gated images at different parts of the cardiac cycle allows the assessment of cardiac function by multi-detector row CT (MDCT) at the time of non-invasive coronary imaging. We compared the accuracy of such measurements by MDCT to cine magnetic resonance (MR). Forty patients underwent the assessment of global and regional cardiac function by 16-slice MDCT and cine MR. Left ventricular (LV) end-diastolic and end-systolic volumes estimated by MDCT (134±51 and 67±56 ml) were similar to those by MR (137±57 and 70±60 ml, respectively; both P=NS) and strongly correlated (r=0.92 and r=0.95, respectively; both P<0.001). Consequently, LV ejection fractions by MDCT and MR were also similar (55±21 vs. 56±21%; P=NS) and highly correlated (r=0.95; P<0.001). Regional end-diastolic and end-systolic wall thicknesses by MDCT were highly correlated (r=0.84 and r=0.92, respectively; both P<0.001), but significantly lower than by MR (8.3±1.8 vs. 8.8±1.9 mm and 12.7±3.4 vs. 13.3±3.5 mm, respectively; both P<0.001). Values of regional wall thickening by MDCT and MR were similar (54±30 vs. 51±31%; P=NS) and also correlated well (r=0.91; P<0.001). Retrospectively gated MDCT can accurately estimate LV volumes, EF and regional LV wall thickening compared to cine MR. (orig.)

  18. Qualitative analysis of intracranial CSF flow on cine-MR imaging, with special reference to signal ratio of CSF to fat tissue

    International Nuclear Information System (INIS)

    Kadowaki, Chikafusa; Hara, Mitsuhiro; Numoto, Mitsuo; Takeuchi, Kazuo; Saito, Isamu

    1993-01-01

    Cine magnetic resonance images (MR) dramatically demonstrate the pulsatile flow of cerebrospinal fluid (CSF) stimulated by the pulsatile motion of the brain following cardiac pulsation. Reduced signal intensity, frequently observed especially in the aqueduct of Sylvius, the third ventricle and the fourth ventricle, is believed to reflect the pulsatile motion of the CSF. Qualitative analysis of MR signal intensity of CSF on each cine frame is compared with CSF flow within the ventricles on real-time cine MR images. While the chronological changes in signal intensities of CSF within the ventricles show only marginal changes in signal intensity in the third ventricle related to downward flow of CSF passing through the foramen of Monro during the early stage of cardiac systole, these changes are thought to have no significant correlation with the CSF flow in the CSF pathway. The chronological changes in relative signal ratios, SR [signal intensities of CSF/signal intensities of fat] can show CSF flow and turbulence within the ventricles. Under normal conditions, within the third ventricle the SR decreases due to pulsatile CSF flow through the foramen of Monro during the early stage of cardiac systole, and decreases because of the flow of CSF from the anterior to the posterior part of the third ventricle, the downward flow of CSF through the aqueduct leads to a lower SR during cardiac diastole. These changes in the fourth ventricle are stimulated by the changes in SR in the third ventricle. The new method of analyzing chronological changes in the relative MR signal ratio of CSF to fat [SR] has the distinct advantage of providing an accurate evaluation of CSF dynamics, and it provides us with important diagnostic information leading to clarification of the pathophysiology of CSF dynamics. (author)

  19. Dosimetry in radiotherapy using a-Si EPIDs: Systems, methods, and applications focusing on 3D patient dose estimation

    Science.gov (United States)

    McCurdy, B. M. C.

    2013-06-01

    An overview is provided of the use of amorphous silicon electronic portal imaging devices (EPIDs) for dosimetric purposes in radiation therapy, focusing on 3D patient dose estimation. EPIDs were originally developed to provide on-treatment radiological imaging to assist with patient setup, but there has also been a natural interest in using them as dosimeters since they use the megavoltage therapy beam to form images. The current generation of clinically available EPID technology, amorphous-silicon (a-Si) flat panel imagers, possess many characteristics that make them much better suited to dosimetric applications than earlier EPID technologies. Features such as linearity with dose/dose rate, high spatial resolution, realtime capability, minimal optical glare, and digital operation combine with the convenience of a compact, retractable detector system directly mounted on the linear accelerator to provide a system that is well-suited to dosimetric applications. This review will discuss clinically available a-Si EPID systems, highlighting dosimetric characteristics and remaining limitations. Methods for using EPIDs in dosimetry applications will be discussed. Dosimetric applications using a-Si EPIDs to estimate three-dimensional dose in the patient during treatment will be overviewed. Clinics throughout the world are implementing increasingly complex treatments such as dynamic intensity modulated radiation therapy and volumetric modulated arc therapy, as well as specialized treatment techniques using large doses per fraction and short treatment courses (ie. hypofractionation and stereotactic radiosurgery). These factors drive the continued strong interest in using EPIDs as dosimeters for patient treatment verification.

  20. Automatic Prostate Tracking and Motion Assessment in Volumetric Modulated Arc Therapy With an Electronic Portal Imaging Device

    International Nuclear Information System (INIS)

    Azcona, Juan Diego; Li, Ruijiang; Mok, Edward; Hancock, Steven; Xing, Lei

    2013-01-01

    Purpose: To assess the prostate intrafraction motion in volumetric modulated arc therapy treatments using cine megavoltage (MV) images acquired with an electronic portal imaging device (EPID). Methods and Materials: Ten prostate cancer patients were treated with volumetric modulated arc therapy using a Varian TrueBeam linear accelerator equipped with an EPID for acquiring cine MV images during treatment. Cine MV images acquisition was scheduled for single or multiple treatment fractions (between 1 and 8). A novel automatic fiducial detection algorithm that can handle irregular multileaf collimator apertures, field edges, fast leaf and gantry movement, and MV image noise and artifacts in patient anatomy was used. All sets of images (approximately 25,000 images in total) were analyzed to measure the positioning accuracy of implanted fiducial markers and assess the prostate movement. Results: Prostate motion can vary greatly in magnitude among different patients. Different motion patterns were identified, showing its unpredictability. The mean displacement and standard deviation of the intrafraction motion was generally less than 2.0 ± 2.0 mm in each of the spatial directions. In certain patients, however, the percentage of the treatment time in which the prostate is displaced more than 5 mm from its planned position in at least 1 spatial direction was 10% or more. The maximum prostate displacement observed was 13.3 mm. Conclusion: Prostate tracking and motion assessment was performed with MV imaging and an EPID. The amount of prostate motion observed suggests that patients will benefit from its real-time monitoring. Megavoltage imaging can provide the basis for real-time prostate tracking using conventional linear accelerators

  1. Free-breathing Sparse Sampling Cine MR Imaging with Iterative Reconstruction for the Assessment of Left Ventricular Function and Mass at 3.0 T.

    Science.gov (United States)

    Sudarski, Sonja; Henzler, Thomas; Haubenreisser, Holger; Dösch, Christina; Zenge, Michael O; Schmidt, Michaela; Nadar, Mariappan S; Borggrefe, Martin; Schoenberg, Stefan O; Papavassiliu, Theano

    2017-01-01

    Purpose To prospectively evaluate the accuracy of left ventricle (LV) analysis with a two-dimensional real-time cine true fast imaging with steady-state precession (trueFISP) magnetic resonance (MR) imaging sequence featuring sparse data sampling with iterative reconstruction (SSIR) performed with and without breath-hold (BH) commands at 3.0 T. Materials and Methods Ten control subjects (mean age, 35 years; range, 25-56 years) and 60 patients scheduled to undergo a routine cardiac examination that included LV analysis (mean age, 58 years; range, 20-86 years) underwent a fully sampled segmented multiple BH cine sequence (standard of reference) and a prototype undersampled SSIR sequence performed during a single BH and during free breathing (non-BH imaging). Quantitative analysis of LV function and mass was performed. Linear regression, Bland-Altman analysis, and paired t testing were performed. Results Similar to the results in control subjects, analysis of the 60 patients showed excellent correlation with the standard of reference for single-BH SSIR (r = 0.93-0.99) and non-BH SSIR (r = 0.92-0.98) for LV ejection fraction (EF), volume, and mass (P 3.0 T is noninferior to the standard of reference irrespective of BH commands. LV mass, however, is overestimated with SSIR. © RSNA, 2016 Online supplemental material is available for this article.

  2. SIFT-based dense pixel tracking on 0.35 T cine-MR images acquired during image-guided radiation therapy with application to gating optimization

    Energy Technology Data Exchange (ETDEWEB)

    Mazur, Thomas R., E-mail: tmazur@radonc.wustl.edu, E-mail: hli@radonc.wustl.edu; Fischer-Valuck, Benjamin W.; Wang, Yuhe; Yang, Deshan; Mutic, Sasa; Li, H. Harold, E-mail: tmazur@radonc.wustl.edu, E-mail: hli@radonc.wustl.edu [Department of Radiation Oncology, Washington University School of Medicine, 4921 Parkview Place, Campus Box 8224, St. Louis, Missouri 63110 (United States)

    2016-01-15

    Purpose: To first demonstrate the viability of applying an image processing technique for tracking regions on low-contrast cine-MR images acquired during image-guided radiation therapy, and then outline a scheme that uses tracking data for optimizing gating results in a patient-specific manner. Methods: A first-generation MR-IGRT system—treating patients since January 2014—integrates a 0.35 T MR scanner into an annular gantry consisting of three independent Co-60 sources. Obtaining adequate frame rates for capturing relevant patient motion across large fields-of-view currently requires coarse in-plane spatial resolution. This study initially (1) investigate the feasibility of rapidly tracking dense pixel correspondences across single, sagittal plane images (with both moderate signal-to-noise and spatial resolution) using a matching objective for highly descriptive vectors called scale-invariant feature transform (SIFT) descriptors associated to all pixels that describe intensity gradients in local regions around each pixel. To more accurately track features, (2) harmonic analysis was then applied to all pixel trajectories within a region-of-interest across a short training period. In particular, the procedure adjusts the motion of outlying trajectories whose relative spectral power within a frequency bandwidth consistent with respiration (or another form of periodic motion) does not exceed a threshold value that is manually specified following the training period. To evaluate the tracking reliability after applying this correction, conventional metrics—including Dice similarity coefficients (DSCs), mean tracking errors (MTEs), and Hausdorff distances (HD)—were used to compare target segmentations obtained via tracking to manually delineated segmentations. Upon confirming the viability of this descriptor-based procedure for reliably tracking features, the study (3) outlines a scheme for optimizing gating parameters—including relative target position and a

  3. Investigation of the dosimetric properties of an a-Si flat panel epid

    International Nuclear Information System (INIS)

    Fielding, A.L.; Jahangir, S.T.

    2004-01-01

    Full text: Electronic portal imaging devices (EPIDs) are primarily used as an electronic replacement for film to verify the set-up of radiotherapy patients based on imaged anatomy. There has recently been much interest in the use of amorphous silicon (a-Si) flat panel EPIDs for dosimetric verification in radiotherapy. The work presented here has been carried out to determine their suitability for dosimetric applications by investigating some of the basic response characteristics and the implications these might have. The measurements reported in this paper were performed using 6-MV photon beams from an Elekta Precise linear accelerator fitted with Elekta iViewGT amorphous silicon flat panel EPIDs. Measurements were performed to investigate the response of the EPID as a function of exposure and field size. Similar measurements were made with an ionisation chamber for comparison. Further measurements were carried out to investigate the response of the EPID to multiple low dose exposures (e.g. 5x2 MU) such as might be encountered in Intensity Modulated Radiotherapy (IMRT). This was compared with the response to a single high dose exposure (e.g. 10 MU) and repeated for a range of exposures. The results show the response of the EPID, to a good approximation, to be linear with dose over the range of 1 -200 MU. However, 'under-responses' in the EPID of up to 5% were seen at the lowest exposures. For multiple low dose segments the sum of the EPID responses was found to be less than the response to the same total exposure in a single large segment. This effect reduces with increase in the magnitude of the low dose segments. The variation in EPID response with field size was found to be greater than that indicated by the ionisation chamber. The results show that the a-Si detector responds to dose, to a good approximation, in a linear manner. The EPID under-response at low doses is thought to be related to the so called ghosting effect. Each image frame has a residual

  4. Accuracy and effectiveness of self-gating signals in free-breathing three-dimensional cardiac cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Li Shuo; Gao Song; Wang Lei; Zhu Yan-Chun; Yang Jie; Xie Yao-Qin; Fu Nan; Wang Yi

    2016-01-01

    Conventional multiple breath-hold two-dimensional (2D) balanced steady-state free precession (SSFP) presents many difficulties in cardiac cine magnetic resonance imaging (MRI). Recently, a self-gated free-breathing three-dimensional (3D) SSFP technique has been proposed as an alternative in many studies. However, the accuracy and effectiveness of self-gating signals have been barely studied before. Since self-gating signals are crucially important in image reconstruction, a systematic study of self-gating signals and comparison with external monitored signals are needed.Previously developed self-gated free-breathing 3D SSFP techniques are used on twenty-eight healthy volunteers. Both electrocardiographic (ECG) and respiratory bellow signals are also acquired during the scan as external signals. Self-gating signal and external signal are compared by trigger and gating window. Gating window is proposed to evaluate the accuracy and effectiveness of respiratory self-gating signal. Relative deviation of the trigger and root-mean-square-deviation of the cycle duration are calculated. A two-tailed paired t-test is used to identify the difference between self-gating and external signals. A Wilcoxon signed rank test is used to identify the difference between peak and valley self-gating triggers.The results demonstrate an excellent correlation ( P = 0, R > 0.99) between self-gating and external triggers. Wilcoxon signed rank test shows that there is no significant difference between peak and valley self-gating triggers for both cardiac ( H = 0, P > 0.10) and respiratory ( H = 0, P > 0.44) motions. The difference between self-gating and externally monitored signals is not significant (two-tailed paired-sample t-test: H = 0, P > 0.90).The self-gating signals could demonstrate cardiac and respiratory motion accurately and effectively as ECG and respiratory bellow. The difference between the two methods is not significant and can be explained. Furthermore, few ECG trigger errors

  5. Quality considerations on cine-imaging and PTCA-fluoroscopy anticipating a digital future

    International Nuclear Information System (INIS)

    Leeuw, P. de

    1986-01-01

    In a modern catheterization laboratory coronary cineangiography, PTCA procedures and digital radiography are performed with one and the same X-ray system. On the basis of an optimization analysis of the image quality using the concepts of window signal-to-noise ratio and equivalent blur, overall performance can roughly be estimated. Some important aspects of a realistic X-ray system design resulting from this analysis have been identified. Specifically, the X-ray loadability and its loading strategy play a crucial role with respect to signal detection sensitivity and the safe, efficient use of X-ray radiation. The analysis shows also that some basic limitations exist to the use of digital subtraction techniques for moving objects. Last but not least, it shows that the video camera performance is critical with respect to the imaging tasks during PTCA and digital procedures. (Auth.)

  6. Trial of quantitative analysis of cardiac function by 3D reconstruction of multislice cine MR images

    International Nuclear Information System (INIS)

    Yamamoto, Hideki; Sei, Tetsurou; Nakagawa, Tomio; Hiraki, Yoshio.

    1994-01-01

    Non-invasive techniques for measuring the dynamic behavior of the left ventricle (LV) can be invaluable tool in the diagnosis of the heart disease. In this paper we present methods for quantitative analysis of cardiac function using a compact magnetic resonance image processing system. A 256 x 256 magnetic resonance transaxial image of the left ventricle in a normal case is obtained. After gray level thresholding and region segmentation, the boundary of the left ventricular chamber is extracted. Then, the boundaries of the left ventricular chamber are displayed three-dimensionally by using the Z-buffer algorithm. Thus, LV volume and ejection fraction are calculated. Here, the value of LV ejection fraction is 60%. These results agree reasonably well with the corresponding data obtained by the echocardiography. (author)

  7. Assessment of an amorphous silicon EPID for quality assurance of enhanced dynamic wedge

    International Nuclear Information System (INIS)

    Greer, P.

    2004-01-01

    Full text: Routine quality assurance (QA) of enhanced dynamic wedge (EDW) is usually performed weekly to monthly. Wedge factors are measured with ion-chamber, and profiles usually with diode-arrays such as the Profiler. The use of an electronic portal imaging device (EPID) for these measurements would combine these into a single rapid set-up and measurement. Currently the Varian EPID in standard imaging mode will not acquire integrated images during EDW treatments, and therefore has not been utilised for EDW dosimetry. Modification to image acquisition was made to enable imaging for EDW, and the performance of the EPID for suitability for quality assurance of EDW was investigated. The accuracy of EDW profiles measured with the EPID were assessed by comparison to Profiler measurements. The EPID was positioned at 105 cm to the detector surface, with 4 cm of additional solid water build-up to give total build-up including EPID inherent build-up of 5 cm. Images of EDW fields were acquired with continuous frame-averaging throughout the delivery. Field sizes of 10x10 cm, and 20x20 cm were used for 30 deg and 60 deg wedge angles for both 6 MV and 18 MV x-rays. Profiler measurements of the same fields were made with 5 cm of solid water build-up with 105 cm to the detector. Profiles in the wedged direction along the central axis of the beam were then compared. The reproducibility of the EPID measured profiles was assessed by three measurements made at weekly intervals. The accuracy of EPID measured wedge factors was investigated with the same experimental set-up. Three images of a 10x10 cm open field were acquired, and the mean pixel value in a 9x9 pixel region at the central axis was found. As the pixel value is the average of all acquired frames, this was multiplied by the number of frames to yield an integrated pixel value. This was repeated for three 10x10 cm 60 deg wedge irradiations. The wedge factor measured with the EPID was then compared to routine weekly

  8. Clinical validation of an in-house EPID dosimetry system for IMRT QA at the Prince of Wales Hospital

    Science.gov (United States)

    Tyler, M.; Vial, P.; Metcalfe, P.; Downes, S.

    2013-06-01

    In this study a simple method using standard flood-field corrected Electronic Portal Imaging Device (EPID) images for routine Intensity Modulated Radiation Therapy (IMRT) Quality Assurance (QA) was investigated. The EPID QA system was designed and tested on a Siemens Oncor Impression linear accelerator with an OptiVue 1000ST EPID panel (Siemens Medical Solutions USA, Inc, USA) and an Elekta Axesse linear accelerator with an iViewGT EPID (Elekta AB, Sweden) for 6 and 10 MV IMRT fields with Step-and-Shoot and dynamic-MLC delivery. Two different planning systems were used for patient IMRT field generation for comparison with the measured EPID fluences. All measured IMRT plans had >95% agreement to the planning fluences (using 3 cGy / 3 mm Gamma Criteria) and were comparable to the pass-rates calculated using a 2-D diode array dosimeter.

  9. Clinical validation of an in-house EPID dosimetry system for IMRT QA at the Prince of Wales Hospital

    International Nuclear Information System (INIS)

    Tyler, M; Downes, S; Vial, P; Metcalfe, P

    2013-01-01

    In this study a simple method using standard flood-field corrected Electronic Portal Imaging Device (EPID) images for routine Intensity Modulated Radiation Therapy (IMRT) Quality Assurance (QA) was investigated. The EPID QA system was designed and tested on a Siemens Oncor Impression linear accelerator with an OptiVue 1000ST EPID panel (Siemens Medical Solutions USA, Inc, USA) and an Elekta Axesse linear accelerator with an iViewGT EPID (Elekta AB, Sweden) for 6 and 10 MV IMRT fields with Step-and-Shoot and dynamic-MLC delivery. Two different planning systems were used for patient IMRT field generation for comparison with the measured EPID fluences. All measured IMRT plans had >95% agreement to the planning fluences (using 3 cGy / 3 mm Gamma Criteria) and were comparable to the pass-rates calculated using a 2-D diode array dosimeter.

  10. Initial observations using a novel "cine" magnetic resonance imaging technique to detect changes in abdominal motion caused by encapsulating peritoneal sclerosis.

    Science.gov (United States)

    Wright, Benjamin; Summers, Angela; Fenner, John; Gillott, Richard; Hutchinson, Charles E; Spencer, Paul A; Wilkie, Martin; Hurst, Helen; Herrick, Sarah; Brenchley, Paul; Augustine, Titus; Bardhan, Karna D

    2011-01-01

    Encapsulating peritoneal sclerosis (EPS) is an uncommon complication of peritoneal dialysis (PD), with high mortality and morbidity. The peritoneum thickens, dysfunctions, and forms a cocoon that progressively "strangulates" the small intestine, causing malnutrition, ischemia, and infarction. There is as yet no reliable noninvasive means of diagnosis, but recent developments in image analysis of cine magnetic resonance imaging for the recognition of adhesions offers a way forward. We used this protocol before surgery in 3 patients with suspected EPS. Image analysis revealed patterns of abdominal movement that were markedly different from the patterns in healthy volunteers. The volunteers showed marked movement throughout the abdomen; in contrast, movement in EPS patients was restricted to just below the diaphragm. This clear difference provides early "proof of principle" of the approach that we have developed.

  11. Vortex Flow in the Right Atrium Surrogates Supraventricular Arrhythmia and Thrombus After Atriopulmonary Connection-Type Fontan Operation: Vortex Flow Analysis Using Conventional Cine Magnetic Resonance Imaging.

    Science.gov (United States)

    Shiina, Yumi; Inai, Kei; Takahashi, Tatsunori; Shimomiya, Yamato; Ishizaki, Umiko; Fukushima, Kenji; Nagao, Michinobu

    2018-02-01

    We developed a novel imaging technique, designated as vortex flow (VF) mapping, which presents a vortex flow visually on conventional two-dimensional (2D) cine MRI. Using it, we assessed circumferential VF patterns and influences on RA thrombus and supraventricular tachycardia (SVT) in AP connection-type Fontan circulation. Retrospectively, we enrolled 27 consecutive patients (25.1 ± 9.2 years) and 7 age-matched controls who underwent cardiac MRI. Conventional cine images acquired using a 1.5-Tesla scanner were scanned for axial and coronal cross section of the RA. We developed "vortex flow mapping" to demonstrate the ratio of the circumferential voxel movement at each phase to the total movement throughout a cardiac cycle towards the RA center. The maximum ratio was used as a magnitude of vortex flow (MVF%) in RA cine imaging. We also measured percentages of strong and weak VF areas (VFA%). Furthermore, in 10 out of 27, we compared VF between previous CMR (3.8 ± 1.5 years ago) and latest CMR. Of the patients, 15 had cardiovascular complications (Group A); 12 did not (Group B). A transaxial image showed that strong VFA% in Group A was significantly smaller than that in Group B or controls. A coronal view revealed that strong VFA% was also smaller, and weak VFA% was larger in Group A than in Group B or controls (P < 0.05, and P < 0.05). Maximum MVF% in Group A was significantly smaller than in other groups (P < 0.001). Univariate logistic analyses revealed weak VFA% on a coronal image, and serum total bilirubin level as factors affecting cardiovascular complications (Odds ratio 1.14 and 66.1, 95% CI 1.004-1.30 and 1.59-2755.6, P values < 0.05 and < 0.05, respectively). Compared to the previous CMR, smaller maximum VMF%, smaller strong VFA%, and larger weak VFA% were identified in the latest CMR. Circumferentially weak VFA% on a coronal image can be one surrogate marker of SVT and thrombus in AP connection-type Fontan circulation. This simple VF

  12. El silencio del monstruo: imagénes espectrales en el cine de terror americano de comienzos del sonoro / Silence of the Monster. Ghostly Images in American Horror Film at the Beginning of Sound

    Directory of Open Access Journals (Sweden)

    Roberto Cueto

    2016-09-01

    Full Text Available Drácula y El doctor Frankenstein inauguraron en 1931 la producción de cine de terror en el cine sonoro al presentar a dos de los iconos por excelencia del género. Sin embargo, al ser películas producidas en el período de transición del cine mudo al sonoro, hacen gala de una característica especial que las diferencia notablemente de la posterior producción terrorífica de Hollywood: la presentación de la monstruosidad sin el recurso de la música extradiegética, lo que genera en la narración un efecto de «agujero» de silencio. En posteriores producciones de cine de terror se intentará paliar estos lapsos de silencio con el empleo de la música extradiegética o ciertos efectos de sonido, lo que parece indicar que, para la norma del cine clásico americano, eran defectos que debían ser corregidos. El siguiente ensayo tiene como objeto subrayar cómo esa llamativa presencia del silencio confiere a la representación de la monstruosidad cierta cualidad de «imágenes espectrales» o «cuerpos siniestros» que el cine de Hollywood prefirió evitar. Por otra parte, se analizará también por qué fue tan importante para el cine clásico americano el empleo de la música extradiegética, ya que cumplía una serie de funciones necesarias para sus estrategias narrativas: una mayor implicación del espectador, una adscripción a los códigos del género terrorífico y un efecto de «tridimensionalidad» aplicado a las imágenes bidimensionales.Palabras clave: música cinematográfica, cine de terror, sonido cinematográfico, transición del cine mudo al sonoro.AbstractDracula and Frankenstein, both released in 1931, inaugurated the production of horror films in Hollywood when they presented the two greatest icons of the genre. As films produced in the transitional period from silent to sound film, both showcase special features that distinguished them from posterior horror films in Hollywood: the introduction of the monstrosity without

  13. Determination of left and right ventricular ejection and filling by fast cine MR imaging in the breath-hold technique in healthy subjects

    International Nuclear Information System (INIS)

    Rominger, M.B.; Bachmann, G.F.; Geuer, M.; Puzik, M.; Rau, W.S.; Pabst, W.

    1999-01-01

    Purpose: Evaluation and comparison of localized and global left and right ventricular ejection and filling with fast cine MR imaging in the breath-hold technique. Materials and Methods: 10 healthy volunteers were examined with a 1.5 Tesla unit and phased-array-coil using a segmented FLASH-2D sequence in breath-hold technique. Peak ejection and peak filling rates [PER, PFR enddiastolic volume (EDV)/s], time to PER and PFR [TPER, TPFR ms] and time of endsystole [TSYS in % RR-intervall] of all slices (complete-slice-evaluation) were evaluated and compared to three left ventricular and one right ventricular slices (reduced three-slice-evaluation). Results: There were significant regional left ventricular differences of PER (p=0.002) and PFR (p=0.007), but not of TPER and TPFR. Ejection and filling indices of the left ventricular middle slice were closest to the overall evaluation of all sections. In the left-/right-side comparison the right ventricular PFR was higher than the left ventricular (5.1 and 4.2 EDV/s) and the right ventricular TPFR was earlier than the left (92.2 and 123.5 ms). Conclusions: With fast cine techniques, regional and global left and right ventricular ejection and filling indices can be evaluated in addition to the global heart volume indices. The three-slice-evaluation represents a comprehensive, clear and time-saving method for daily routine. (orig.) [de

  14. Monte Carlo modelling of a-Si EPID response: The effect of spectral variations with field size and position

    International Nuclear Information System (INIS)

    Parent, Laure; Seco, Joao; Evans, Phil M.; Fielding, Andrew; Dance, David R.

    2006-01-01

    This study focused on predicting the electronic portal imaging device (EPID) image of intensity modulated radiation treatment (IMRT) fields in the absence of attenuation material in the beam with Monte Carlo methods. As IMRT treatments consist of a series of segments of various sizes that are not always delivered on the central axis, large spectral variations may be observed between the segments. The effect of these spectral variations on the EPID response was studied with fields of various sizes and off-axis positions. A detailed description of the EPID was implemented in a Monte Carlo model. The EPID model was validated by comparing the EPID output factors for field sizes between 1x1 and 26x26 cm 2 at the isocenter. The Monte Carlo simulations agreed with the measurements to within 1.5%. The Monte Carlo model succeeded in predicting the EPID response at the center of the fields of various sizes and offsets to within 1% of the measurements. Large variations (up to 29%) of the EPID response were observed between the various offsets. The EPID response increased with field size and with field offset for most cases. The Monte Carlo model was then used to predict the image of a simple test IMRT field delivered on the beam axis and with an offset. A variation of EPID response up to 28% was found between the on- and off-axis delivery. Finally, two clinical IMRT fields were simulated and compared to the measurements. For all IMRT fields, simulations and measurements agreed within 3%--0.2 cm for 98% of the pixels. The spectral variations were quantified by extracting from the spectra at the center of the fields the total photon yield (Y total ), the photon yield below 1 MeV (Y low ), and the percentage of photons below 1 MeV (P low ). For the studied cases, a correlation was shown between the EPID response variation and Y total , Y low , and P low

  15. A methodology to accurately quantify patellofemoral cartilage contact kinematics by combining 3D image shape registration and cine-PC MRI velocity data.

    Science.gov (United States)

    Borotikar, Bhushan S; Sipprell, William H; Wible, Emily E; Sheehan, Frances T

    2012-04-05

    Patellofemoral osteoarthritis and its potential precursor patellofemoral pain syndrome (PFPS) are common, costly, and debilitating diseases. PFPS has been shown to be associated with altered patellofemoral joint mechanics; however, an actual variation in joint contact stresses has not been established due to challenges in accurately quantifying in vivo contact kinematics (area and location). This study developed and validated a method for tracking dynamic, in vivo cartilage contact kinematics by combining three magnetic resonance imaging (MRI) techniques, cine-phase contrast (CPC), multi-plane cine (MPC), and 3D high-resolution static imaging. CPC and MPC data were acquired from 12 healthy volunteers while they actively extended/flexed their knee within the MRI scanner. Since no gold standard exists for the quantification of in vivo dynamic cartilage contact kinematics, the accuracy of tracking a single point (patellar origin relative to the femur) represented the accuracy of tracking the kinematics of an entire surface. The accuracy was determined by the average absolute error between the PF kinematics derived through registration of MPC images to a static model and those derived through integration of the CPC velocity data. The accuracy ranged from 0.47 mm to 0.77 mm for the patella and femur and from 0.68 mm to 0.86 mm for the patellofemoral joint. For purely quantifying joint kinematics, CPC remains an analytically simpler and more accurate (accuracy <0.33 mm) technique. However, for application requiring the tracking of an entire surface, such as quantifying cartilage contact kinematics, this combined imaging approach produces accurate results with minimal operator intervention. Published by Elsevier Ltd.

  16. Characterization of the a-Si EPID in the unity MR-linac for dosimetric applications

    Science.gov (United States)

    Torres-Xirau, I.; Olaciregui-Ruiz, I.; Baldvinsson, G.; Mijnheer, B. J.; van der Heide, U. A.; Mans, A.

    2018-01-01

    Electronic portal imaging devices (EPIDs) are frequently used in external beam radiation therapy for dose verification purposes. The aim of this study was to investigate the dose-response characteristics of the EPID in the Unity MR-linac (Elekta AB, Stockholm, Sweden) relevant for dosimetric applications under clinical conditions. EPID images and ionization chamber (IC) measurements were used to study the effects of the magnetic field, the scatter generated in the MR housing reaching the EPID, and inhomogeneous attenuation from the MR housing. Dose linearity and dose rate dependencies were also determined. The magnetic field strength at EPID level did not exceed 10 mT, and dose linearity and dose rate dependencies proved to be comparable to that on a conventional linac. Profiles of fields, delivered with and without the magnetic field, were indistinguishable. The EPID center had an offset of 5.6 cm in the longitudinal direction, compared to the beam central axis, meaning that large fields in this direction will partially fall outside the detector area and not be suitable for verification. Beam attenuation by the MRI scanner and the table is gantry angle dependent, presenting a minimum attenuation of 67% relative to the 90° measurement. Repeatability, observed over two months, was within 0.5% (1 SD). In order to use the EPID for dosimetric applications in the MR-linac, challenges related to the EPID position, scatter from the MR housing, and the inhomogeneous, gantry angle-dependent attenuation of the beam will need to be solved.

  17. Accuracy and effectiveness of self-gating signals in free-breathing three-dimensional cardiac cine magnetic resonance imaging

    Science.gov (United States)

    Li, Shuo; Wang, Lei; Zhu, Yan-Chun; Yang, Jie; Xie, Yao-Qin; Fu, Nan; Wang, Yi; Gao, Song

    2016-12-01

    Conventional multiple breath-hold two-dimensional (2D) balanced steady-state free precession (SSFP) presents many difficulties in cardiac cine magnetic resonance imaging (MRI). Recently, a self-gated free-breathing three-dimensional (3D) SSFP technique has been proposed as an alternative in many studies. However, the accuracy and effectiveness of self-gating signals have been barely studied before. Since self-gating signals are crucially important in image reconstruction, a systematic study of self-gating signals and comparison with external monitored signals are needed. Previously developed self-gated free-breathing 3D SSFP techniques are used on twenty-eight healthy volunteers. Both electrocardiographic (ECG) and respiratory bellow signals are also acquired during the scan as external signals. Self-gating signal and external signal are compared by trigger and gating window. Gating window is proposed to evaluate the accuracy and effectiveness of respiratory self-gating signal. Relative deviation of the trigger and root-mean-square-deviation of the cycle duration are calculated. A two-tailed paired t-test is used to identify the difference between self-gating and external signals. A Wilcoxon signed rank test is used to identify the difference between peak and valley self-gating triggers. The results demonstrate an excellent correlation (P = 0, R > 0.99) between self-gating and external triggers. Wilcoxon signed rank test shows that there is no significant difference between peak and valley self-gating triggers for both cardiac (H = 0, P > 0.10) and respiratory (H = 0, P > 0.44) motions. The difference between self-gating and externally monitored signals is not significant (two-tailed paired-sample t-test: H = 0, P > 0.90). The self-gating signals could demonstrate cardiac and respiratory motion accurately and effectively as ECG and respiratory bellow. The difference between the two methods is not significant and can be explained. Furthermore, few ECG trigger errors

  18. Control of an X-ray cine radiography apparatus

    International Nuclear Information System (INIS)

    Nishio, K.

    1982-01-01

    This patent application describes an X-ray cine radiography apparatus comprising an X-ray tube, an image intensifier for converting the X-rays transmitted through an object into a visual image and a cine camera for picking up the visual image, a photomultiplier detects the brightness of the visual image to produce a brightness signal and a potentiometer detects the actual tube voltage of said X-ray tube. (author)

  19. Physiological Uptake of 18F-Fluorodeoxyglucose in Uterine Endometrium and Myometrium: Correlation with Uterine Motility Evaluated by Cine Magnetic Resonance Imaging

    International Nuclear Information System (INIS)

    Kido, A.; Nishizawa, S.; Okada, H.; Nakamoto, Y.; Yamamoto, A.; Fujimoto, K.; Togashi, K.

    2009-01-01

    Background: Accumulation of 18 F-fluorodeoxyglucose ( 18 F-FDG) in the uterine endometrium and uterine motility are dependent on menstrual cycle. However, the relationship between them remains unknown. Purpose: To investigate the relationship between radiometabolic activity of 18 F-FDG in the uterus and uterine motility observed by cine magnetic resonance imaging (MRI). Material and Methods: The study population consisted of 65 healthy, fertile women, selected from 229 women who underwent positron emission tomography (PET), computed tomography (CT), and MRI for cancer screening at our facility. They were divided into three groups according to their menstrual cycle phases: menstrual, follicular-periovulatory, and luteal. Regions of interest (ROIs) were placed over the endometrium and myometrium to calculate the standardized uptake value (SUV). Uterine peristalsis and contraction shown by cine MR imaging were evaluated visually, and the correlation between FDG uptake and uterine movements was assessed. Results: After excluding nine patients due to inadequate images, 56 patients (19 follicular-periovulatory, 27 luteal, and 10 menstrual) were analyzed. FDG uptake of the endometrium, frequency of peristalsis, and the presence of sustained contraction varied according to the menstruation cycle, with a tendency toward greater uptake in the menstrual phase, but there was little relationship between the frequency of uterine peristalsis and FDG accumulation in the uterus. Significantly higher FDG accumulation in the endometrium was observed in patients with sustained contractions (3.32±1.47) than in those without contractions (2.45±0.66). Conclusion: Our preliminary data suggest that FDG accumulation in the endometrium tends to be higher in patients with uterine contraction, although there was no significant correlation between uterine peristalsis and FDG uptake in the uterine myometrium or endometrium

  20. Physiological Uptake of 18F-Fluorodeoxyglucose in Uterine Endometrium and Myometrium: Correlation with Uterine Motility Evaluated by Cine Magnetic Resonance Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Kido, A.; Nishizawa, S.; Okada, H. (Hamamatsu Medical Imaging Center, Hamamatsu Medical Photonics Foundation, Hamakita City, Shizuoka (Japan)); Nakamoto, Y.; Yamamoto, A.; Fujimoto, K.; Togashi, K. Dept. of Diagnostic Radiology, Kyoto Univ. Hospital, Kyoto City, Kyoto (Japan))

    2009-05-15

    Background: Accumulation of 18F-fluorodeoxyglucose (18F-FDG) in the uterine endometrium and uterine motility are dependent on menstrual cycle. However, the relationship between them remains unknown. Purpose: To investigate the relationship between radiometabolic activity of 18F-FDG in the uterus and uterine motility observed by cine magnetic resonance imaging (MRI). Material and Methods: The study population consisted of 65 healthy, fertile women, selected from 229 women who underwent positron emission tomography (PET), computed tomography (CT), and MRI for cancer screening at our facility. They were divided into three groups according to their menstrual cycle phases: menstrual, follicular-periovulatory, and luteal. Regions of interest (ROIs) were placed over the endometrium and myometrium to calculate the standardized uptake value (SUV). Uterine peristalsis and contraction shown by cine MR imaging were evaluated visually, and the correlation between FDG uptake and uterine movements was assessed. Results: After excluding nine patients due to inadequate images, 56 patients (19 follicular-periovulatory, 27 luteal, and 10 menstrual) were analyzed. FDG uptake of the endometrium, frequency of peristalsis, and the presence of sustained contraction varied according to the menstruation cycle, with a tendency toward greater uptake in the menstrual phase, but there was little relationship between the frequency of uterine peristalsis and FDG accumulation in the uterus. Significantly higher FDG accumulation in the endometrium was observed in patients with sustained contractions (3.32+-1.47) than in those without contractions (2.45+-0.66). Conclusion: Our preliminary data suggest that FDG accumulation in the endometrium tends to be higher in patients with uterine contraction, although there was no significant correlation between uterine peristalsis and FDG uptake in the uterine myometrium or endometrium

  1. Comparison of left ventricular function assessment using phonocardiogram- and electrocardiogram-triggered 2D SSFP CINE MR imaging at 1.5 T and 3.0 T

    Energy Technology Data Exchange (ETDEWEB)

    Becker, Meike [University Hospital, RWTH Aachen, Department of Diagnostic Radiology, Aachen (Germany); Humboldt-University, Experimental and Clinical Research Center (ECRC), Charite Campus Buch, Berlin (Germany); Frauenrath, Tobias; Hezel, Fabian [University Hospital, RWTH Aachen, Department of Diagnostic Radiology, Aachen (Germany); Berlin Ultrahigh Field Facility, Max-Delbrueck Center for Molecular Medicine, Berlin (Germany); Krombach, Gabriele A.; Kremer, Ute; Koppers, Benedikt [University Hospital, RWTH Aachen, Department of Diagnostic Radiology, Aachen (Germany); Butenweg, Christoph; Goemmel, Andreas [Chair of Structural Statics and Dynamics, RWTH Aachen, Aachen (Germany); Utting, Jane F. [MRI, NHS Grampian, Aberdeen Royal Infirmary, Aberdeen (United Kingdom); Schulz-Menger, Jeanette [Humboldt-University, Working Group Cardiovascular MR, Franz-Volhard-Klinik, Department of Cardiology, HELIOS-Klinikum Berlin-Buch and Charite Campus Buch, Berlin (Germany); Niendorf, Thoralf [University Hospital, RWTH Aachen, Department of Diagnostic Radiology, Aachen (Germany); Humboldt-University, Experimental and Clinical Research Center (ECRC), Charite Campus Buch, Berlin (Germany); Berlin Ultrahigh Field Facility, Max-Delbrueck Center for Molecular Medicine, Berlin (Germany)

    2010-06-15

    As high-field cardiac MRI (CMR) becomes more widespread the propensity of ECG to interference from electromagnetic fields (EMF) and to magneto-hydrodynamic (MHD) effects increases and with it the motivation for a CMR triggering alternative. This study explores the suitability of acoustic cardiac triggering (ACT) for left ventricular (LV) function assessment in healthy subjects (n = 14). Quantitative analysis of 2D CINE steady-state free precession (SSFP) images was conducted to compare ACT's performance with vector ECG (VCG). Endocardial border sharpness (EBS) was examined paralleled by quantitative LV function assessment. Unlike VCG, ACT provided signal traces free of interference from EMF or MHD effects. In the case of correct R-wave recognition, VCG-triggered 2D CINE SSFP was immune to cardiac motion effects - even at 3.0 T. However, VCG-triggered 2D SSFP CINE imaging was prone to cardiac motion and EBS degradation if R-wave misregistration occurred. ACT-triggered acquisitions yielded LV parameters (end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF) and left ventricular mass (LVM)) comparable with those derived from VCG-triggered acquisitions (1.5 T: ESV{sub VCG} = (56 {+-} 17) ml, EDV{sub VCG} = (151 {+-} 32) ml, LVM{sub VCG} = (97 {+-} 27) g, SV{sub VCG} = (94 {+-} 19) ml, EF{sub VCG} = (63 {+-} 5)% cf. ESV{sub ACT} = (56 {+-} 18) ml, EDV{sub ACT} = (147 {+-} 36) ml, LVM{sub ACT} = (102 {+-} 29) g, SV{sub ACT} = (91 {+-} 22) ml, EF{sub ACT} = (62 {+-} 6)%; 3.0 T: ESV{sub VCG} = (55 {+-} 21) ml, EDV{sub VCG} = (151 {+-} 32) ml, LVM{sub VCG} = (101 {+-} 27) g, SV{sub VCG} = (96 {+-} 15) ml, EF{sub VCG} = (65 {+-} 7)% cf. ESV{sub ACT} = (54 {+-} 20) ml, EDV{sub ACT} = (146 {+-} 35) ml, LVM{sub ACT} = (101 {+-} 30) g, SV{sub ACT} = (92 {+-} 17) ml, EF{sub ACT} = (64 {+-} 6)%). ACT's intrinsic insensitivity to interference from electromagnetic fields renders it suitable for clinical CMR. (orig.)

  2. Application of cine MRI-tagging method to aortic dessection

    International Nuclear Information System (INIS)

    Yoshioka, Kunihiro; Takahashi, Tsuneo; Kamata, Hiroyuki; Kikuchi, Kenichi; Yamaguchi, Kojiro.

    1992-01-01

    For the evaluation of aortic dissection, ECG-gated SE or cine MR imaging has been usually performed. However, detection of slow flow in the false lumen and differentiation between mural thrombus and slow flow are sometimes difficult. Because paradoxical enhancement due to slow blood flow simulates thrombus. We performed cine MR imaging with persaturation tagging, which clearly showed differentiation between thrombus and paradoxical enhancement. We concluded that cine MR imaging with tagging method was useful to evaluate the slow flow and thrombus in the false lumen. (author)

  3. Image-guided radiotherapy in near real time with intensity-modulated radiotherapy megavoltage treatment beam imaging.

    Science.gov (United States)

    Mao, Weihua; Hsu, Annie; Riaz, Nadeem; Lee, Louis; Wiersma, Rodney; Luxton, Gary; King, Christopher; Xing, Lei; Solberg, Timothy

    2009-10-01

    To utilize image-guided radiotherapy (IGRT) in near real time by obtaining and evaluating the online positions of implanted fiducials from continuous electronic portal imaging device (EPID) imaging of prostate intensity-modulated radiotherapy (IMRT) delivery. Upon initial setup using two orthogonal images, the three-dimensional (3D) positions of all implanted fiducial markers are obtained, and their expected two-dimensional (2D) locations in the beam's-eye-view (BEV) projection are calculated for each treatment field. During IMRT beam delivery, EPID images of the megavoltage treatment beam are acquired in cine mode and subsequently analyzed to locate 2D locations of fiducials in the BEV. Simultaneously, 3D positions are estimated according to the current EPID image, information from the setup portal images, and images acquired at other gantry angles (the completed treatment fields). The measured 2D and 3D positions of each fiducial are compared with their expected 2D and 3D setup positions, respectively. Any displacements larger than a predefined tolerance may cause the treatment system to suspend the beam delivery and direct the therapists to reposition the patient. Phantom studies indicate that the accuracy of 2D BEV and 3D tracking are better than 1 mm and 1.4 mm, respectively. A total of 7330 images from prostate treatments were acquired and analyzed, showing a maximum 2D displacement of 6.7 mm and a maximum 3D displacement of 6.9 mm over 34 fractions. This EPID-based, real-time IGRT method can be implemented on any external beam machine with portal imaging capabilities without purchasing any additional equipment, and there is no extra dose delivered to the patient.

  4. Image-Guided Radiotherapy in Near Real Time With Intensity-Modulated Radiotherapy Megavoltage Treatment Beam Imaging

    International Nuclear Information System (INIS)

    Mao Weihua; Hsu, Annie; Riaz, Nadeem; Lee, Louis; Wiersma, Rodney; Luxton, Gary; King, Christopher; Xing Lei; Solberg, Timothy

    2009-01-01

    Purpose: To utilize image-guided radiotherapy (IGRT) in near real time by obtaining and evaluating the online positions of implanted fiducials from continuous electronic portal imaging device (EPID) imaging of prostate intensity-modulated radiotherapy (IMRT) delivery. Methods and Materials: Upon initial setup using two orthogonal images, the three-dimensional (3D) positions of all implanted fiducial markers are obtained, and their expected two-dimensional (2D) locations in the beam's-eye-view (BEV) projection are calculated for each treatment field. During IMRT beam delivery, EPID images of the megavoltage treatment beam are acquired in cine mode and subsequently analyzed to locate 2D locations of fiducials in the BEV. Simultaneously, 3D positions are estimated according to the current EPID image, information from the setup portal images, and images acquired at other gantry angles (the completed treatment fields). The measured 2D and 3D positions of each fiducial are compared with their expected 2D and 3D setup positions, respectively. Any displacements larger than a predefined tolerance may cause the treatment system to suspend the beam delivery and direct the therapists to reposition the patient. Results: Phantom studies indicate that the accuracy of 2D BEV and 3D tracking are better than 1 mm and 1.4 mm, respectively. A total of 7330 images from prostate treatments were acquired and analyzed, showing a maximum 2D displacement of 6.7 mm and a maximum 3D displacement of 6.9 mm over 34 fractions. Conclusions: This EPID-based, real-time IGRT method can be implemented on any external beam machine with portal imaging capabilities without purchasing any additional equipment, and there is no extra dose delivered to the patient.

  5. Dosimetric characterization of an a-based EPID for quality control if patient-specific IMRT

    International Nuclear Information System (INIS)

    Larrinaga Cortina, Eduardo Francisco; Alfonso Laguardia, Rodolfo; Silvestre Patallo, Ileana; Garcia Yip, Fernando

    2009-01-01

    The Electronic portal imaging devices, EPID for its acronym in English is a technology widely used for verification of patient positioning on linear accelerators routinely. Its use as a dosimetry device is not as widespread, although many researches in this field. It assessed the availability and versatility of the use EPID based on an amorphous silicon (a-Si) as a means of quality control specific patient for a methodology of Radiation Intensity Modulated IMRT. Dosimetric parameters were determined for the linearity of dose versus response, dispersion and sensitivity factors off-axis radiation. For absolute measurements the linearity of the dose-response relationship EPID was better than 1.1 and 1.5% for photon beams of 6 and 15mV respectively, in the range from 2 to 500 UM. The dose dependence with field size was studied and compared with the factors of dispersion in water at different depths, in agreement with those measured at 5 cm depth, Scp (z = 5cm). Off-axis sensitivity of the EPID was determined by comparing the measured profiles versus the same profiles at different depths in water. The best correspondence was observed at 5 cm depth, where the EPID response underestimates the dose to 4% for all sizes of fields in the plateau area. The EPID can be used for the evaluation of dosimetric parameters of the beam at a specific depth in water of 5 cm and a discrepancy in an acceptable maximum rate of 4%. (author)

  6. Highly-Accelerated Real-Time Cardiac Cine MRI Using k-t SPARSE-SENSE

    Science.gov (United States)

    Feng, Li; Srichai, Monvadi B.; Lim, Ruth P.; Harrison, Alexis; King, Wilson; Adluru, Ganesh; Dibella, Edward VR.; Sodickson, Daniel K.; Otazo, Ricardo; Kim, Daniel

    2012-01-01

    For patients with impaired breath-hold capacity and/or arrhythmias, real-time cine MRI may be more clinically useful than breath-hold cine MRI. However, commercially available real-time cine MRI methods using parallel imaging typically yield relatively poor spatio-temporal resolution due to their low image acquisition speed. We sought to achieve relatively high spatial resolution (~2.5mm × 2.5mm) and temporal resolution (~40ms), to produce high-quality real-time cine MR images that could be applied clinically for wall motion assessment and measurement of left ventricular (LV) function. In this work, we present an 8-fold accelerated real-time cardiac cine MRI pulse sequence using a combination of compressed sensing and parallel imaging (k-t SPARSE-SENSE). Compared with reference, breath-hold cine MRI, our 8-fold accelerated real-time cine MRI produced significantly worse qualitative grades (1–5 scale), but its image quality and temporal fidelity scores were above 3.0 (adequate) and artifacts and noise scores were below 3.0 (moderate), suggesting that acceptable diagnostic image quality can be achieved. Additionally, both 8-fold accelerated real-time cine and breath-hold cine MRI yielded comparable LV function measurements, with coefficient of variation cine MRI with k-t SPARSE-SENSE is a promising modality for rapid imaging of myocardial function. PMID:22887290

  7. Highly accelerated real-time cardiac cine MRI using k-t SPARSE-SENSE.

    Science.gov (United States)

    Feng, Li; Srichai, Monvadi B; Lim, Ruth P; Harrison, Alexis; King, Wilson; Adluru, Ganesh; Dibella, Edward V R; Sodickson, Daniel K; Otazo, Ricardo; Kim, Daniel

    2013-07-01

    For patients with impaired breath-hold capacity and/or arrhythmias, real-time cine MRI may be more clinically useful than breath-hold cine MRI. However, commercially available real-time cine MRI methods using parallel imaging typically yield relatively poor spatio-temporal resolution due to their low image acquisition speed. We sought to achieve relatively high spatial resolution (∼2.5 × 2.5 mm(2)) and temporal resolution (∼40 ms), to produce high-quality real-time cine MR images that could be applied clinically for wall motion assessment and measurement of left ventricular function. In this work, we present an eightfold accelerated real-time cardiac cine MRI pulse sequence using a combination of compressed sensing and parallel imaging (k-t SPARSE-SENSE). Compared with reference, breath-hold cine MRI, our eightfold accelerated real-time cine MRI produced significantly worse qualitative grades (1-5 scale), but its image quality and temporal fidelity scores were above 3.0 (adequate) and artifacts and noise scores were below 3.0 (moderate), suggesting that acceptable diagnostic image quality can be achieved. Additionally, both eightfold accelerated real-time cine and breath-hold cine MRI yielded comparable left ventricular function measurements, with coefficient of variation cine MRI with k-t SPARSE-SENSE is a promising modality for rapid imaging of myocardial function. Copyright © 2012 Wiley Periodicals, Inc.

  8. Age- and gender-specific differences in left and right ventricular cardiac function and mass determined by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Sandstede, J.; Lipke, C.; Beer, M.; Hofmann, S.; Pabst, T.; Kenn, W.; Hahn, D.; Neubauer, S.

    2000-01-01

    We examined possible age- and gender-specific differences in the function and mass of left (LV) and right (RV) ventricles in 36 healthy volunteers using cine gradient-recalled echo magnetic resonance imaging. Subjects were divided into four groups (nine men and nine women in each): men aged under 45 years (32 ± 7), women aged under 45 (27 ± 6), men aged over 45 (59 ± 8), and women aged over 45 (57 ± 9). Functional analysis of cardiac volume and mass and of LV wall motion was performed by manual segmentation of the endocardial and epicardial borders of the end-diastolic and end-systolic frame; both absolute and normalized (per square meter body surface area) values were evaluated. With age there was a significant decrease in both absolute and normalized LV and RV chamber volumes (EDV, ESV), while LV and RV masses remained unchanged. Gender-specific differences were found in cardiac mass and volume (for men and women, respectively: LV mass, 155 ± 18 and 110 ± 16 g; LV EDV, 118 ± 27 and 96 ± 21 ml; LV ESV, 40 ± 13 and 29 ± 9 ml; RV mass, 52 ± 10 and 39 ± 5 g; RV EDV, 131 ± 28 and 100 ± 23 ml; RV ESV, 53 ± 17 and 33 ± 15 ml). Normalization to body surface area eliminated differences in LV volumes but not those in LV mass, RV mass, or RV function. Functional parameters such as cardiac output and LV ejection fraction showed nonsignificant or only slight differences and were thus largely independent of age and gender. Intra- and interobserver variability ranged between 1.4 % and 5.9 % for all parameters. Cine magnetic resonance imaging thus shows age- and gender-specific differences in cardiac function, and therefore the evaluation of cardiac function in patients should consider age- and gender-matched normative values. (orig.)

  9. Integrative analysis of cardiac function and metabolism in patients with idiopathic hypertrophic cardiomyopathy with H-1 cine MR imaging and P-31 MR spectroscopy

    International Nuclear Information System (INIS)

    Wagner, S.; Chew, W.M.; Semelka, R.; Tomei, E.; Caputo, G.; O'Sullivan, M.; Chatterjee, K.; Parmley, W.W.; Wolfe, C.L.; Higgins, C.B.

    1989-01-01

    The purpose of the study was to use MR imaging and P-31 spectroscopy for the functional and metabolic characterization of patients with hypertrophic cardiomyopathy (HCM). Nine patients with HCM underwent combined P-31 spectroscopy (one-dimensional chemical shift imaging) and MR imaging at 1.5 T. MR imaging localized the distribution of ventricular hypertrophy and cine MR imaging quantitated cardiac volumes, contractility, and left ventricular mass. The phosphodiester/phosphocreatine ratio (PDE/PCr) in six HCM patients was not different from normal, but in three patients it was significantly (P <.01) higher. This finding could not be attributed to contractible abnormalities, the distribution of hypertrophy, or the left ventricular mass. The abnormal PDE/PCr ratio was found only in patients with echocardiographic findings of resting left ventricular outflow obstruction but was not associated with severity or distribution of hypertrophy. The study shows the use of combined MR imaging and MR spectroscopy to characterize HCM and possibly to identify abnormal myocardial tissue

  10. Cine-MR imaging aqueductal CSF flow in normal pressure hydrocephalus syndrome before and after CSF shunt

    International Nuclear Information System (INIS)

    Mascalchi, M.; Arnetoli, G.; Inzitari, D.; Dal Pozzo, G.; Lolli, F.; Caramella, D.; Bartolozzi, C.

    1993-01-01

    Reproducibility of the aqueductal CSF signal intensity on a gradient echo cine-MR sequence exploiting through plane inflow enhancement was tested in 11 patients with normal or dilated ventricles. Seven patients with normal pressure hydrocephalus (NPH) syndrome were investigated with the sequence before and after CSF shunting. Two patients exhibiting central flow void within a hyperintense aqueductal CSF improved after surgery and the flow void disappeared after shunting. One patient with increased maximum and minimum aqueductal CSF signal as compared to 18 healthy controls also improved and the aqueductal CSF signal was considerably decreased after shunting. Three patients with aqueductal CSF values similar to those in the controls did not improve, notwithstanding their maximum aqueductal CSF signals decreasing slightly after shunting. No appreciable aqueductal CSF flow related enhancement consistent with non-communicating hydrocephalus was found in the last NPH patient who improved after surgery. Cine-MR with inflow technique yields a reproducible evaluation of flow-related aqueductal CSF signal changes which might help in identifying shunt responsive NPH patients. These are likely to be those with hyperdynamic aqueductal CSF or aqueductal obstruction. (orig.)

  11. Música y Cine

    OpenAIRE

    Olarte Martínez, Matilde María

    2008-01-01

    Materiales de clase: Introducción a la Historia de la Música en el Cine; partituras de cine mudo; música clásica y cine; textos sobre música incidental. Se Analiza la música utilizada en cinematografía, y el estudio de sus funciones aplicada a la imagen. Audiciones de música incidental y música preexistente de los principales compositores. la música utilizada en cinematografía, y el estudio de sus funciones aplicada a la imagen. Audiciones de música incidental y música preexistente de los ...

  12. Managing the backscatter component from the robotic arm of an a-Si EPID

    International Nuclear Information System (INIS)

    Lee, C.G.; Menk, F.; Greer, P.B.

    2010-01-01

    Full text: Backscatter from the robotic arm mechanism of an a-Si EPID in IMRT images was examined. Images corrected with a conventional flood field (FF) containing a backscatter component (BSC) from the robotic ann were compared with a BSC-free FF. A Yarian 21 EX linac (6 MV, 18 MV) was used. All images were acquired with two aS500 EPIDs, one R-arm and one E-arm. The BSC of an EPID image is the ratio of an image acquired with the EPID attached to the arm then detaching the arm from the EPID and acquiring the same image. A range of square field sizes from 2.5 x 2.5 cm to 27.5 x 27.5 cm were acquired and the BSC analyzed. The BSC of the FFs were also measured. A series of IMRT fields were acquired. Each field was corrected with a conventional FF and compared with a BSC-free FF. Figure I shows the magnitude of the BSC from each arm in the inplane for a 6 x beam. Square fields above 16 x l6 cm (R-arm) and lO x 10 cm (E-arm) benefited from a conventional FF as it tended to cancel out the BSC in the acquired square field. The opposite was observed for smaller field sizes. A gamma analysis of the IMRT fields showed a FF correction containing a BSC reduces the effect of the arm in the final image. IMRT EPID images using conventional FFs have been shown to be less affected by backscatter from the robotic arm compared to BSC-free flood fields. (author)

  13. Dual breath-hold magnetic resonance cine evaluation of global and regional cardiac function

    International Nuclear Information System (INIS)

    Wintersperger, Bernd J.; Dietrich, Olaf; Huber, Armin; Reiser, Maximilian F.; Schoenberg, Stefan O.; Sincleair, Spencer; Runge, Val M.

    2007-01-01

    The purpose of our study was to evaluate the accuracy of a multislice cine magnetic resonance imaging (MRI) technique with parallel imaging in regard to global and regional left ventricular function. Forty-two individuals underwent cine MRI on a 1.5-tesla scanner. Cine MRI used a steady-state free precession technique and was performed as a single-slice technique (nonTSENSE cine) and an accelerated multislice technique (TSENSE cine) with five slices per breath-hold. End diastolic volume (EDV), end systolic volume (ESV), and ejection fraction (EF) were evaluated for all data sets and in regard to regional wall motion and regional wall motion analysis, and quantitative regional wall thickness and systolic thickening were also assessed. EDV, ESV, and EF based on TSENSE cine showed excellent correlation to the nonTSENSE cine approach (all r 2 =0.99, P<0.001). While EDV evaluations showed a small underestimation for TSENSE cine, ESV and EF showed accurate results compared with nonTSENSE cine. Both readers showed good agreement (κ=0.72) in regional wall motion assessment comparing both techniques. Data acquisition for the multislice approach was significantly shorter (∝75%) that in single-slice cine. We conclude that accurate evaluation of regional wall motion and left ventricular EF is possible using accelerated multislice cine MR with high spatial and temporal resolution. (orig.)

  14. Contrast-enhanced specific absorption rate-efficient 3D cardiac cine with respiratory-triggered radiofrequency gating.

    Science.gov (United States)

    Henningsson, Markus; Chan, Raymond H; Goddu, Beth; Goepfert, Lois A; Razavi, Reza; Botnar, Rene M; Schaeffter, Tobias; Nezafat, Reza

    2013-04-01

    To investigate the use of radiofrequency (RF) gating in conjunction with a paramagnetic contrast agent to reduce the specific absorption rate (SAR) and increase the blood-myocardium contrast in balanced steady-state free precession (bSSFP) 3D cardiac cine. RF gating was implemented by synchronizing the RF-excitation with an external respiratory sensor (bellows), which could additionally be used for respiratory gating. For reference, respiratory-gated 3D cine images were acquired without RF gating. Free-breathing 3D cine images were acquired in eight healthy subjects before and after contrast injection (Gd-BOPTA) and compared to breath-hold 2D cine. RF-gated 3D cine reduced the SAR by nearly 40% without introducing significant artifacts while providing left ventricle (LV) measurements similar to those obtained with 2D cine. The contrast-to-noise ratio (CNR) was significantly higher for 3D cine compared to 2D cine, both before and after contrast injection; however, no statistically significant CNR increase was observed for the postcontrast 3D cine compared to the precontrast acquisitions. Respiratory-triggered RF gating significantly reduces SAR in 3D cine acquisitions, which may enable a more widespread clinical use of 3D cine. Furthermore, CNR of 3D bSSFP cine is higher than of 2D and administration of Gd-BOPTA does not improve the CNR of 3D cine. Copyright © 2012 Wiley Periodicals, Inc.

  15. Implementation of DMLC quality control using EPID (Portal Dosimetry); Implementacao de um controle de qualidade de DMLC utilizando um EPID (Portal Dosimetry)

    Energy Technology Data Exchange (ETDEWEB)

    Mattos, Fabio R.; Furnari, Laura, E-mail: mattos.fr@gmail.com [Universidade de Sao Paulo (USP), Sao Paulo, SP (Brazil). Faculdade de Medicina; Universidade de Sao Paulo (INRAD/HC/FMUSP), Sao Paulo, SP (Brazil). Instituto de Radiologia. Setor de Radioterapia

    2017-11-01

    A Quality Assurance (QA) to ensure the expected performance of a Multileaf Collimator System (MLC) is essential to deliver dose in a safety and appropriate way. The time required for equipment control and dosimetry may be reduced when the Electronic Portal Image Device (EPID) is used. The aim of this work was to check the resolution limits of the detection system for IMRT mode, and to propose a set of tests that can provide positioning analysis of a multileaf system. A Varian iX Clinac equipped with an 80 leaf Millenium MLC, and an amorphous silicon based EPID (aS1000) was used. The EPID proved itself effective for detecting errors up to 0.5 mm. The proposed tests provided relevant results of leaf position, and revealed that the MLC system is within acceptable limits found in literature. (author)

  16. SU-F-T-476: Performance of the AS1200 EPID for Periodic Photon Quality Assurance

    Energy Technology Data Exchange (ETDEWEB)

    DeMarco, J; Fraass, B; Yang, W; McKenzie Boehnke, E [Cedars-Sinai Medical Center, Los Angeles, CA (United States); Moran, J [University Michigan Medical Center, Ann Arbor, MI (United States); Barnes, M [Calvary Mater Hospital Newcastle, Warratah, NSW (Australia); Greer, P [Calvary Mater Newcastle, Newcastle (Australia); Kim, G [University of California, San Diego, La Jolla, CA (United States)

    2016-06-15

    Purpose: To assess the dosimetric performance of a new amorphous silicon flat-panel electronic portal imaging device (EPID) suitable for high-intensity, flattening-filter-free delivery mode. Methods: An EPID-based QA suite was created with automation to periodically monitor photon central-axis output and two-dimensional beam profile constancy as a function of gantry angle and dose-rate. A Varian TrueBeamTM linear accelerator installed with Developer Mode was used to customize and deliver XML script routines for the QA suite using the dosimetry mode image acquisition for an aS1200 EPID. Automatic post-processing software was developed to analyze the resulting DICOM images. Results: The EPID was used to monitor photon beam output constancy (central-axis), flatness, and symmetry over a period of 10 months for four photon beam energies (6x, 15x, 6xFFF, and 10xFFF). EPID results were consistent to those measured with a standard daily QA check device. At the four cardinal gantry angles, the standard deviation of the EPID central-axis output was <0.5%. Likewise, EPID measurements were independent for the wide range of dose rates (including up to 2400 mu/min for 10xFFF) studied with a standard deviation of <0.8% relative to the nominal dose rate for each energy. Also, profile constancy and field size measurements showed good agreement with the reference acquisition of 0° gantry angle and nominal dose rate. XML script files were also tested for MU linearity and picket-fence delivery. Using Developer Mode, the test suite was delivered in <60 minutes for all 4 photon energies with 4 dose rates per energy and 5 picket-fence acquisitions. Conclusion: Dosimetry image acquisition using a new EPID was found to be accurate for standard and high-intensity photon beams over a broad range of dose rates over 10 months. Developer Mode provided an efficient platform to customize the EPID acquisitions by using custom script files which significantly reduced the time. This work was funded

  17. Phase-contrast cine MR imaging of normal aqueductal CSF flow. Effect of aging and relation to CSF void on modulus MR

    International Nuclear Information System (INIS)

    Barkhof, F.; Kouwenhoven, M.; Scheltens, P.; Sprenger, M.; Algra, P.; Valk, J.

    1994-01-01

    Cine phase-contrast MR imaging was used to study pulsatile CSF flow in the aqueduct in 11 young controls (mean age 30 years) and 9 old controls (mean age 69 years). A high-resolution gradient echo technique and an oblique imaging plane, perpendicular to the aqueduct, was used to avoid volume averaging. Phantom studies confirmed that the technique was accurate. Aqueductal velocity and flux in old controls was higher than in young controls, but the differences were not significant. For all controls together, the averaged peak velocity was 4.2 ± 1.5 cm/s in rostral and -7.8 ± 4.9 cm/s in caudal direction; for the flux it was 0.16 ± 0.10 cm 3 /s in rostral and -0.29 ± 0.19 cm 3 /s in caudal direction. Phase-contrast measurements were significantly related to flow-void on modulus MR images, but not with ventricular size or cortical atrophy. The present technique avoids underestimation of aqueductal flow, and therefore reveals higher aqueductal velocity and flux values than previous studies. Factors other than age or atrophy seem to determine aqueductal CSF flow. (orig.)

  18. Evaluation of highly accelerated real-time cardiac cine MRI in tachycardia.

    Science.gov (United States)

    Bassett, Elwin C; Kholmovski, Eugene G; Wilson, Brent D; DiBella, Edward V R; Dosdall, Derek J; Ranjan, Ravi; McGann, Christopher J; Kim, Daniel

    2014-02-01

    Electrocardiogram (ECG)-gated breath-hold cine MRI is considered to be the gold standard test for the assessment of cardiac function. However, it may fail in patients with arrhythmia, impaired breath-hold capacity and poor ECG gating. Although ungated real-time cine MRI may mitigate these problems, commercially available real-time cine MRI pulse sequences using parallel imaging typically yield relatively poor spatiotemporal resolution because of their low image acquisition efficiency. As an extension of our previous work, the purpose of this study was to evaluate the diagnostic quality and accuracy of eight-fold-accelerated real-time cine MRI with compressed sensing (CS) for the quantification of cardiac function in tachycardia, where it is challenging for real-time cine MRI to provide sufficient spatiotemporal resolution. We evaluated the performances of eight-fold-accelerated cine MRI with CS, three-fold-accelerated real-time cine MRI with temporal generalized autocalibrating partially parallel acquisitions (TGRAPPA) and ECG-gated breath-hold cine MRI in 21 large animals with tachycardia (mean heart rate, 104 beats per minute) at 3T. For each cine MRI method, two expert readers evaluated the diagnostic quality in four categories (image quality, temporal fidelity of wall motion, artifacts and apparent noise) using a Likert scale (1-5, worst to best). One reader evaluated the left ventricular functional parameters. The diagnostic quality scores were significantly different between the three cine pulse sequences, except for the artifact level between CS and TGRAPPA real-time cine MRI. Both ECG-gated breath-hold cine MRI and eight-fold accelerated real-time cine MRI yielded all four scores of ≥ 3.0 (acceptable), whereas three-fold-accelerated real-time cine MRI yielded all scores below 3.0, except for artifact (3.0). The left ventricular ejection fraction (LVEF) measurements agreed better between ECG-gated cine MRI and eight-fold-accelerated real-time cine MRI

  19. Comparison of ghosting effects for three commercial a-Si EPIDs

    International Nuclear Information System (INIS)

    McDermott, L. N.; Nijsten, S. M. J. J. G.; Sonke, J.-J.; Partridge, M.; Herk, M. van; Mijnheer, B. J.

    2006-01-01

    Many studies have reported dosimetric characteristics of amorphous silicon electronic portal imaging devices (EPIDs). Some studies ascribed a non-linear signal to gain ghosting and image lag. Other reports, however, state the effect is negligible. This study compares the signal-to-monitor unit (MU) ratio for three different brands of EPID systems. The signal was measured for a wide range of monitor units (5-1000), dose-rates, and beam energies. All EPIDs exhibited a relative under-response for beams of few MUs; giving 4 to 10% lower signal-to-MU ratios relative to that of 1000 MUs. This under-response is consistent with ghosting effects due to charge trapping

  20. Cine CT in the evaluation of coronary bypass graft patency

    International Nuclear Information System (INIS)

    Stanford, W.; Rooholamini, M.; Rumberger, J.; Marcus, M.; Hiratzka, L.

    1986-01-01

    Cine CT produces axial images over an 8-cm section of the aorta in 50 msec. This characteristic makes the technique useful for evaluating coronary bypass graft (CBG) patency. With the use of 40 ml of 67% iothalamate sodium injected via an antecubital vein, 28 patients with 68 CBGs underwent cine CT. Ten patients with 21 CBGs also underwent cardiac catheterization. In the latter group the overall accuracy of cine CT compared to cardiac catheterization was 95.2% (20/21). The sensitivity was 94.1% (16/17), and the specificity was 100% (4/4). This figure compares favorably with the 92% sensitivity achieved with conventional CT

  1. Time dependent pre-treatment EPID dosimetry for standard and FFF VMAT.

    Science.gov (United States)

    Podesta, Mark; Nijsten, Sebastiaan M J J G; Persoon, Lucas C G G; Scheib, Stefan G; Baltes, Christof; Verhaegen, Frank

    2014-08-21

    Methods to calibrate Megavoltage electronic portal imaging devices (EPIDs) for dosimetry have been previously documented for dynamic treatments such as intensity modulated radiotherapy (IMRT) using flattened beams and typically using integrated fields. While these methods verify the accumulated field shape and dose, the dose rate and differential fields remain unverified. The aim of this work is to provide an accurate calibration model for time dependent pre-treatment dose verification using amorphous silicon (a-Si) EPIDs in volumetric modulated arc therapy (VMAT) for both flattened and flattening filter free (FFF) beams. A general calibration model was created using a Varian TrueBeam accelerator, equipped with an aS1000 EPID, for each photon spectrum 6 MV, 10 MV, 6 MV-FFF, 10 MV-FFF. As planned VMAT treatments use control points (CPs) for optimization, measured images are separated into corresponding time intervals for direct comparison with predictions. The accuracy of the calibration model was determined for a range of treatment conditions. Measured and predicted CP dose images were compared using a time dependent gamma evaluation using criteria (3%, 3 mm, 0.5 sec). Time dependent pre-treatment dose verification is possible without an additional measurement device or phantom, using the on-board EPID. Sufficient data is present in trajectory log files and EPID frame headers to reliably synchronize and resample portal images. For the VMAT plans tested, significantly more deviation is observed when analysed in a time dependent manner for FFF and non-FFF plans than when analysed using only the integrated field. We show EPID-based pre-treatment dose verification can be performed on a CP basis for VMAT plans. This model can measure pre-treatment doses for both flattened and unflattened beams in a time dependent manner which highlights deviations that are missed in integrated field verifications.

  2. On the use of EPID-based implanted marker tracking for 4D radiotherapy

    International Nuclear Information System (INIS)

    Keall, P.J.; Todor, A.D.; Vedam, S.S.; Bartee, C.L.; Siebers, J.V.; Kini, V.R.; Mohan, R.

    2004-01-01

    Four-dimensional (4D) radiotherapy delivery to dynamically moving tumors requires a real-time signal of the tumor position as a function of time so that the radiation beam can continuously track the tumor during the respiration cycle. The aim of this study was to develop and evaluate an electronic portal imaging device (EPID)-based marker-tracking system that can be used for real-time tumor targeting, or 4D radiotherapy. Three gold cylinders, 3 mm in length and 1 mm in diameter, were implanted in a dynamic lung phantom. The phantom range of motion was 4 cm with a 3-s 'breathing' period. EPID image acquisition parameters were modified, allowing image acquisition in 0.1 s. Images of the stationary and moving phantom were acquired. Software was developed to segment automatically the marker positions from the EPID images. Images acquired in 0.1 s displayed higher noise and a lower signal-noise ratio than those obtained using regular (>1 s) acquisition settings. However, the markers were still clearly visible on the 0.1-s images. The motion of the phantom blurred the images of the markers and further reduced the signal-noise ratio, though they could still be successfully segmented from the images in 10-30 ms of computation time. The positions of gold markers placed in the lung phantom were detected successfully, even for phantom velocities substantially higher than those observed for typical lung tumors. This study shows that using EPID-based marker tracking for 4D radiotherapy is feasible, however, changes in linear accelerator technology and EPID-based image acquisition as well as patient studies are required before this method can be implemented clinically

  3. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation.

    Science.gov (United States)

    Rowshanfarzad, P; Riis, H L; Zimmermann, S J; Ebert, M A

    2015-07-01

    In radiotherapy treatments, it is crucial to monitor the performance of linear accelerator (linac) components, including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt and the sag in leaf bank assembly owing to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with 5 ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of the image data. Nine Elekta AB (Stockholm, Sweden) linacs of different models and number of years in service were investigated. The average EPID sag was within 2 mm for all tested linacs. Some machines showed >1-mm gantry sag. Changes in the SDD values were within 1.3 cm. EPID skewness and tilt values were <1° in all machines. The maximum sag in multileaf collimator leaf bank assemblies was around 1 mm. A meaningful correlation was found between the age of the linacs and their mechanical performance. Conclusions and Advances in knowledge: The method and software developed in this study provide a simple tool for effective investigation of the behaviour of Elekta linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Elekta machines.

  4. Left ventricular strain and its pattern estimated from cine CMR and validation with DENSE

    International Nuclear Information System (INIS)

    Gao, Hao; Luo, Xiaoyu; Allan, Andrew; McComb, Christie; Berry, Colin

    2014-01-01

    Measurement of local strain provides insight into the biomechanical significance of viable myocardium. We attempted to estimate myocardial strain from cine cardiovascular magnetic resonance (CMR) images by using a b-spline deformable image registration method. Three healthy volunteers and 41 patients with either recent or chronic myocardial infarction (MI) were studied at 1.5 Tesla with both cine and DENSE CMR. Regional circumferential and radial left ventricular strains were estimated from cine and DENSE acquisitions. In all healthy volunteers, there was no difference for peak circumferential strain (− 0.18 ± 0.04 versus − 0.18 ± 0.03, p = 0.76) between cine and DENSE CMR, however peak radial strain was overestimated from cine (0.84 ± 0.37 versus 0.49 ± 0.2, p < 0.01). In the patient study, the peak strain patterns predicted by cine were similar to the patterns from DENSE, including the strain evolution related to recovery time and strain patterns related to MI scar extent. Furthermore, cine-derived strain disclosed different strain patterns in MI and non-MI regions, and regions with transmural and non-transmural MI as DENSE. Although there were large variations with radial strain measurements from cine CMR images, useful circumferential strain information can be obtained from routine clinical CMR imaging. Cine strain analysis has potential to improve the diagnostic yield from routine CMR imaging in clinical practice. (paper)

  5. Left ventricular strain and its pattern estimated from cine CMR and validation with DENSE.

    Science.gov (United States)

    Gao, Hao; Allan, Andrew; McComb, Christie; Luo, Xiaoyu; Berry, Colin

    2014-07-07

    Measurement of local strain provides insight into the biomechanical significance of viable myocardium. We attempted to estimate myocardial strain from cine cardiovascular magnetic resonance (CMR) images by using a b-spline deformable image registration method. Three healthy volunteers and 41 patients with either recent or chronic myocardial infarction (MI) were studied at 1.5 Tesla with both cine and DENSE CMR. Regional circumferential and radial left ventricular strains were estimated from cine and DENSE acquisitions. In all healthy volunteers, there was no difference for peak circumferential strain (- 0.18 ± 0.04 versus - 0.18 ± 0.03, p = 0.76) between cine and DENSE CMR, however peak radial strain was overestimated from cine (0.84 ± 0.37 versus 0.49 ± 0.2, p cine were similar to the patterns from DENSE, including the strain evolution related to recovery time and strain patterns related to MI scar extent. Furthermore, cine-derived strain disclosed different strain patterns in MI and non-MI regions, and regions with transmural and non-transmural MI as DENSE. Although there were large variations with radial strain measurements from cine CMR images, useful circumferential strain information can be obtained from routine clinical CMR imaging. Cine strain analysis has potential to improve the diagnostic yield from routine CMR imaging in clinical practice.

  6. Automated Motion Estimation for 2D Cine DENSE MRI

    Science.gov (United States)

    Gilliam, Andrew D.; Epstein, Frederick H.

    2013-01-01

    Cine displacement encoding with stimulated echoes (DENSE) is a magnetic resonance (MR) method that directly encodes tissue displacement into MR phase images. This technique has successfully interrogated many forms of tissue motion, but is most commonly used to evaluate cardiac mechanics. Currently, motion analysis from cine DENSE images requires manually delineated anatomical structures. An automated analysis would improve measurement throughput, simplify data interpretation, and potentially access important physiological information during the MR exam. In this article, we present the first fully automated solution for the estimation of tissue motion and strain from 2D cine DENSE data. Results using both simulated and human cardiac cine DENSE data indicate good agreement between the automated algorithm and the standard semi-manual analysis method. PMID:22575669

  7. Comparative performance evaluation of a new a-Si EPID that exceeds quad high-definition resolution.

    Science.gov (United States)

    McConnell, Kristen A; Alexandrian, Ara; Papanikolaou, Niko; Stathakis, Sotiri

    2018-01-01

    Electronic portal imaging devices (EPIDs) are an integral part of the radiation oncology workflow for treatment setup verification. Several commercial EPID implementations are currently available, each with varying capabilities. To standardize performance evaluation, Task Group Report 58 (TG-58) and TG-142 outline specific image quality metrics to be measured. A LinaTech Image Viewing System (IVS), with the highest commercially available pixel matrix (2688x2688 pixels), was independently evaluated and compared to an Elekta iViewGT (1024x1024 pixels) and a Varian aSi-1000 (1024x768 pixels) using a PTW EPID QC Phantom. The IVS, iViewGT, and aSi-1000 were each used to acquire 20 images of the PTW QC Phantom. The QC phantom was placed on the couch and aligned at isocenter. The images were exported and analyzed using the epidSoft image quality assurance (QA) software. The reported metrics were signal linearity, isotropy of signal linearity, signal-tonoise ratio (SNR), low contrast resolution, and high-contrast resolution. These values were compared between the three EPID solutions. Computed metrics demonstrated comparable results between the EPID solutions with the IVS outperforming the aSi-1000 and iViewGT in the low and high-contrast resolution analysis. The performance of three commercial EPID solutions have been quantified, evaluated, and compared using results from the PTW QC Phantom. The IVS outperformed the other panels in low and high-contrast resolution, but to fully realize the benefits of the IVS, the selection of the monitor on which to view the high-resolution images is important to prevent down sampling and visual of resolution.

  8. Evaluation of flow volume and flow patterns in the patent false lumen of chronic aortic dissections using velocity-encoded cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Inoue, Toshihisa; Watanabe, Shigeru; Sakurada, Hideki; Ono, Katsuhiro; Urano, Miharu; Hijikata, Yasuyoshi; Saito, Isao; Masuda, Yoshiaki

    2000-01-01

    In 21 patients with chronic aortic dissections and proven patent false lumens, the flow volume and flow patterns in the patent false lumens was evaluated using velocity-encoded cine magnetic resonance imaging (VENC-MRI) and the relationship between the flow characteristics and aortic enlargement was retrospectively examined. Flow patterns in the false lumen were divided into 3 groups: pattern A with primarily antegrade flow (n=6), pattern R with primarily retrograde flow (n=3), and pattern B with bidirectional flow (n=12). In group A, the rate of flow volume in the false lumen compared to the total flow volume in true and false lumens (%TFV) and the average rate of enlargement of the maximum diameter of the dissected aorta per year (ΔD) were significantly greater than in groups R and B (%TFV: 74.1±0.07 vs 15.2±0.03 vs 11.8±0.04, p<0.01; ΔD: 3.62±0.82 vs 0 vs 0.58±0.15 mm/year, p<0.05, respectively). There was a significant correlation between %TFV and ΔD (r=0.79, p<0.0001). Evaluation of flow volume and flow patterns in the patent false lumen using VENC-MRI may be useful for predicting enlargement of the dissected aorta. (author)

  9. Evaluation of flow volume and flow patterns in the patent false lumen of chronic aortic dissections using velocity-encoded cine magnetic resonance imaging

    Energy Technology Data Exchange (ETDEWEB)

    Inoue, Toshihisa; Watanabe, Shigeru; Sakurada, Hideki; Ono, Katsuhiro; Urano, Miharu; Hijikata, Yasuyoshi; Saito, Isao; Masuda, Yoshiaki [Chiba Univ. (Japan). School of Medicine

    2000-10-01

    In 21 patients with chronic aortic dissections and proven patent false lumens, the flow volume and flow patterns in the patent false lumens was evaluated using velocity-encoded cine magnetic resonance imaging (VENC-MRI) and the relationship between the flow characteristics and aortic enlargement was retrospectively examined. Flow patterns in the false lumen were divided into 3 groups: pattern A with primarily antegrade flow (n=6), pattern R with primarily retrograde flow (n=3), and pattern B with bidirectional flow (n=12). In group A, the rate of flow volume in the false lumen compared to the total flow volume in true and false lumens (%TFV) and the average rate of enlargement of the maximum diameter of the dissected aorta per year ({delta}D) were significantly greater than in groups R and B (%TFV: 74.1{+-}0.07 vs 15.2{+-}0.03 vs 11.8{+-}0.04, p<0.01; {delta}D: 3.62{+-}0.82 vs 0 vs 0.58{+-}0.15 mm/year, p<0.05, respectively). There was a significant correlation between %TFV and {delta}D (r=0.79, p<0.0001). Evaluation of flow volume and flow patterns in the patent false lumen using VENC-MRI may be useful for predicting enlargement of the dissected aorta. (author)

  10. Cine-CT: A new technique in X-ray computed tomography

    International Nuclear Information System (INIS)

    Jaschke, W.; Georgi, M.; Lipton, M.J.; Gould, R.G.

    1988-01-01

    The 'Cine-CT' belongs to a new generation of CT devices permitting not only diagnostics on a morphological basis but also function studies. By means of Cine-CT's electron beam technique, image frequencies of up to 17 images per second and exposure times of 50 msec are possible. (orig.) [de

  11. TH-AB-202-01: Daily Lung Tumor Motion Characterization On EPIDs Using a Markerless Tiling Model

    Energy Technology Data Exchange (ETDEWEB)

    Rozario, T [University of Texas Southwestern Medical Center, Dallas, TX (United States); University of Texas at Dallas, Richardson, TX (United States); Chiu, T; Lu, W; Chen, M; Yan, Y [University of Texas Southwestern Medical Center, Dallas, TX (United States); Bereg, S [University of Texas at Dallas, Richardson, TX (United States); Mao, W [University of Texas Southwestern Medical Center, Dallas, TX (United States); Henry Ford Hospital, Detroit, MI (United States)

    2016-06-15

    Purpose: Tracking lung tumor motion in real time allows for target dose escalation while simultaneously reducing dose to sensitive structures, thus increasing local control without increasing toxicity. We present a novel intra-fractional markerless lung tumor tracking algorithm using MV treatment beam images acquired during treatment delivery. Strong signals superimposed on the tumor significantly reduced the soft tissue resolution; while different imaging modalities involved introduce global imaging discrepancies. This reduced the comparison accuracies. A simple yet elegant Tiling algorithm is reported to overcome the aforementioned issues. Methods: MV treatment beam images were acquired continuously in beam’s eye view (BEV) by an electronic portal imaging device (EPID) during treatment and analyzed to obtain tumor positions on every frame. Every frame of the MV image was simulated by a composite of two components with separate digitally reconstructed radiographs (DRRs): all non-moving structures and the tumor. This Titling algorithm divides the global composite DRR and the corresponding MV projection into sub-images called tiles. Rigid registration is performed independently on tile-pairs in order to improve local soft tissue resolution. This enables the composite DRR to be transformed accurately to match the MV projection and attain a high correlation value through a pixel-based linear transformation. The highest cumulative correlation for all tile-pairs achieved over a user-defined search range indicates the 2-D coordinates of the tumor location on the MV projection. Results: This algorithm was successfully applied to cine-mode BEV images acquired during two SBRT plans delivered five times with different motion patterns to each of two phantoms. Approximately 15000 beam’s eye view images were analyzed and tumor locations were successfully identified on every projection with a maximum/average error of 1.8 mm / 1.0 mm. Conclusion: Despite the presence of

  12. Measurement of cerebral blood flow with two-dimensional cine phase-contrast MR imaging. Evaluation of normal subjects and patients with vertigo

    International Nuclear Information System (INIS)

    Kashimada, Akio; Machida, Kikuo; Honda, Norinari; Mamiya, Toshio; Takahashi, Taku; Kamano, Tsuyoshi; Osada, Hisato

    1995-01-01

    The purpose of this study was to determine whether or not the vertebral flow of patients with vertigo and normal brain magnetic resonance (MR) images was decreased in comparison with normal controls. Cerebral blood flow (CBF) was quantitatively measured by a two-dimensional phase contrast cine MR imaging technique in 24 normal controls (mean age, 38.6 years; range, 12-70) and 23 patients (mean age, 53.7 years; range, 19-76) with a 1.5 Tesla MR imaging unit. Inter-and intraobserver variation in blood flow measurements was small (r=0.970, standard error of the estimate [SEE]=2.9 ml, n=80; r=0.963, SEE=4.6 ml, n=40, respectively), In the normal group, mean summed vertebral flow (171 ml/min, SD=40.6) was significantly less than mean summed carotid flow (523 ml/min, SD=111). Right vertebral flow (80.2 ml/min, SD=30.5) was less than left vertebral flow (91.2 ml/min, SD=38.2), but the difference was not statistically significant (p<0.05), In the 23 patients, although the summed vertebral flows of two patients (63.3, 88.8 ml/min) were significantly less than that of the normal group, mean summed vertebral flow (165 ml/min, SD=59.1) showed no significant difference from that of the normal group (p<0.05). In this study, the majority of patients had normal CBF. This method is clinically useful for estimating CBF. (author)

  13. Measurement of cerebral blood flow with two-dimensional cine phase-contrast MR imaging. Evaluation of normal subjects and patients with vertigo

    Energy Technology Data Exchange (ETDEWEB)

    Kashimada, Akio; Machida, Kikuo; Honda, Norinari; Mamiya, Toshio; Takahashi, Taku; Kamano, Tsuyoshi; Osada, Hisato [Saitama Medical School, Kawagoe (Japan). Saitama Medical Center

    1995-03-01

    The purpose of this study was to determine whether or not the vertebral flow of patients with vertigo and normal brain magnetic resonance (MR) images was decreased in comparison with normal controls. Cerebral blood flow (CBF) was quantitatively measured by a two-dimensional phase contrast cine MR imaging technique in 24 normal controls (mean age, 38.6 years; range, 12-70) and 23 patients (mean age, 53.7 years; range, 19-76) with a 1.5 Tesla MR imaging unit. Inter-and intraobserver variation in blood flow measurements was small (r=0.970, standard error of the estimate [SEE]=2.9 ml, n=80; r=0.963, SEE=4.6 ml, n=40, respectively), In the normal group, mean summed vertebral flow (171 ml/min, SD=40.6) was significantly less than mean summed carotid flow (523 ml/min, SD=111). Right vertebral flow (80.2 ml/min, SD=30.5) was less than left vertebral flow (91.2 ml/min, SD=38.2), but the difference was not statistically significant (p<0.05), In the 23 patients, although the summed vertebral flows of two patients (63.3, 88.8 ml/min) were significantly less than that of the normal group, mean summed vertebral flow (165 ml/min, SD=59.1) showed no significant difference from that of the normal group (p<0.05). In this study, the majority of patients had normal CBF. This method is clinically useful for estimating CBF. (author).

  14. El cine posmoderno

    Directory of Open Access Journals (Sweden)

    Edel Cadena Vargas

    2015-05-01

    Full Text Available En el presente trabajo se trata de dar respuesta a las siguientes preguntas: ¿cuál es el origen del cine posmoderno?, ¿cuáles son los factores que permiten su creación?, ¿a qué sentimiento responde?

  15. SU-F-T-259: GPR Tables for the Estimation of Mid-Plane Dose Using EPID

    International Nuclear Information System (INIS)

    Annamalai, Gopiraj; Watanabe, Yoichi

    2016-01-01

    Purpose: To develop a simple method for estimating the mid-plane dose (MPD) of a patient using Electronic Portal imaging Device (EPID). Methods: A Varian TrueBeam with aSi100 EPID was used in this study. The EPID images were acquired for a 30 cm × 30 cm homogeneous slab phantom and a 30 cm diameter 20 cm thick cylindrical phantom in the continuous dosimetry mode. The acquired EPID images in XIM format were imported into in-house MATLAB program for the data analysis. First, the dosimetric characteristics of EPID were studied for dose-response linearity, dose-rate dependence, and field size dependence. Next, the average pixels values of the EPID images were correlated with the MPD measured by an ionisation chamber for various thicknesses of the slab phantom (8 cm – 30 cm) and for various square field sizes (3×3 cm 2 – 25×25 cm 2 at the isocenter). Look-up tables called as GPR tables were then generated for both SSD and SAD setup by taking the ratio of MPD measured by the ionisation chamber and the corresponding EPID pixel values. The accuracy of the GPR tables was evaluated by varying the field size, phantom thickness, and wedge angles with the slab and cylindrical phantoms. Results: The dose response of EPID was linear from 20 MU to 300 MU. The EPID response for different dose rates from 40 MU/min to 600 MU/min was within ±1%. The difference in the doses from the GPR tables and the doses measured by the ionization chambers were within 2% for slab phantoms, and 3% for the cylindrical phantom for various field sizes, phantom thickness, and wedge angles. Conclusion: GPR tables are a ready reckoner for in-vivo dosimetry and it can be used to quickly estimate the MPD value from the EPID images with an accuracy of ±3% for common clinical treatment. project work funded by Union for International cancer control(UICC) under ICRETT fellowship

  16. SU-F-T-259: GPR Tables for the Estimation of Mid-Plane Dose Using EPID

    Energy Technology Data Exchange (ETDEWEB)

    Annamalai, Gopiraj [Government Arignar Anna Memorial Cancer Hospital & Research Institute, Kanchipuram, TAMILNADU (India); Watanabe, Yoichi [University of Minnesota, Minneapolis, MN (United States)

    2016-06-15

    Purpose: To develop a simple method for estimating the mid-plane dose (MPD) of a patient using Electronic Portal imaging Device (EPID). Methods: A Varian TrueBeam with aSi100 EPID was used in this study. The EPID images were acquired for a 30 cm × 30 cm homogeneous slab phantom and a 30 cm diameter 20 cm thick cylindrical phantom in the continuous dosimetry mode. The acquired EPID images in XIM format were imported into in-house MATLAB program for the data analysis. First, the dosimetric characteristics of EPID were studied for dose-response linearity, dose-rate dependence, and field size dependence. Next, the average pixels values of the EPID images were correlated with the MPD measured by an ionisation chamber for various thicknesses of the slab phantom (8 cm – 30 cm) and for various square field sizes (3×3 cm{sup 2} – 25×25 cm{sup 2} at the isocenter). Look-up tables called as GPR tables were then generated for both SSD and SAD setup by taking the ratio of MPD measured by the ionisation chamber and the corresponding EPID pixel values. The accuracy of the GPR tables was evaluated by varying the field size, phantom thickness, and wedge angles with the slab and cylindrical phantoms. Results: The dose response of EPID was linear from 20 MU to 300 MU. The EPID response for different dose rates from 40 MU/min to 600 MU/min was within ±1%. The difference in the doses from the GPR tables and the doses measured by the ionization chambers were within 2% for slab phantoms, and 3% for the cylindrical phantom for various field sizes, phantom thickness, and wedge angles. Conclusion: GPR tables are a ready reckoner for in-vivo dosimetry and it can be used to quickly estimate the MPD value from the EPID images with an accuracy of ±3% for common clinical treatment. project work funded by Union for International cancer control(UICC) under ICRETT fellowship.

  17. Normalize the response of EPID in pursuit of linear accelerator dosimetry standardization.

    Science.gov (United States)

    Cai, Bin; Goddu, S Murty; Yaddanapudi, Sridhar; Caruthers, Douglas; Wen, Jie; Noel, Camille; Mutic, Sasa; Sun, Baozhou

    2018-01-01

    Normalize the response of electronic portal imaging device (EPID) is the first step toward an EPID-based standardization of Linear Accelerator (linac) dosimetry quality assurance. In this study, we described an approach to generate two-dimensional (2D) pixel sensitivity maps (PSM) for EPIDs response normalization utilizing an alternative beam and dark-field (ABDF) image acquisition technique and large overlapping field irradiations. The automated image acquisition was performed by XML-controlled machine operation and the PSM was generated based on a recursive calculation algorithm for Varian linacs equipped with aS1000 and aS1200 imager panels. Cross-comparisons of normalized beam profiles and 1.5%/1.5 mm 1D Gamma analysis was adopted to quantify the improvement of beam profile matching before and after PSM corrections. PSMs were derived for both photon (6, 10, 15 MV) and electron (6, 20 MeV) beams via proposed method. The PSM-corrected images reproduced a horn-shaped profile for photon beams and a relative uniform profiles for electrons. For dosimetrically matched linacs equipped with aS1000 panels, PSM-corrected images showed increased 1D-Gamma passing rates for all energies, with an average 10.5% improvement for crossline and 37% for inline beam profiles. Similar improvements in the phantom study were observed with a maximum improvement of 32% for 15 MV and 22% for 20 MeV. The PSM value showed no significant change for all energies over a 3-month period. In conclusion, the proposed approach correct EPID response for both aS1000 and aS1200 panels. This strategy enables the possibility to standardize linac dosimetry QA and to benchmark linac performance utilizing EPID as the common detector. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  18. Source position verification and dosimetry in HDR brachytherapy using an EPID

    International Nuclear Information System (INIS)

    Smith, R. L.; Taylor, M. L.; McDermott, L. N.; Franich, R. D.; Haworth, A.; Millar, J. L.

    2013-01-01

    Purpose: Accurate treatment delivery in high dose rate (HDR) brachytherapy requires correct source dwell positions and dwell times to be administered relative to each other and to the surrounding anatomy. Treatment delivery inaccuracies predominantly occur for two reasons: (i) anatomical movement or (ii) as a result of human errors that are usually related to incorrect implementation of the planned treatment. Electronic portal imaging devices (EPIDs) were originally developed for patient position verification in external beam radiotherapy and their application has been extended to provide dosimetric information. The authors have characterized the response of an EPID for use with an 192 Ir brachytherapy source to demonstrate its use as a verification device, providing both source position and dosimetric information.Methods: Characterization of the EPID response using an 192 Ir brachytherapy source included investigations of reproducibility, linearity with dose rate, photon energy dependence, and charge build-up effects associated with exposure time and image acquisition time. Source position resolution in three dimensions was determined. To illustrate treatment verification, a simple treatment plan was delivered to a phantom and the measured EPID dose distribution compared with the planned dose.Results: The mean absolute source position error in the plane parallel to the EPID, for dwells measured at 50, 100, and 150 mm source to detector distances (SDD), was determined to be 0.26 mm. The resolution of the z coordinate (perpendicular distance from detector plane) is SDD dependent with 95% confidence intervals of ±0.1, ±0.5, and ±2.0 mm at SDDs of 50, 100, and 150 mm, respectively. The response of the EPID is highly linear to dose rate. The EPID exhibits an over-response to low energy incident photons and this nonlinearity is incorporated into the dose calibration procedure. A distance (spectral) dependent dose rate calibration procedure has been developed. The

  19. Improved attenuation correction for respiratory gated PET/CT with extended-duration cine CT: a simulation study

    Science.gov (United States)

    Zhang, Ruoqiao; Alessio, Adam M.; Pierce, Larry A.; Byrd, Darrin W.; Lee, Tzu-Cheng; De Man, Bruno; Kinahan, Paul E.

    2017-03-01

    Due to the wide variability of intra-patient respiratory motion patterns, traditional short-duration cine CT used in respiratory gated PET/CT may be insufficient to match the PET scan data, resulting in suboptimal attenuation correction that eventually compromises the PET quantitative accuracy. Thus, extending the duration of cine CT can be beneficial to address this data mismatch issue. In this work, we propose to use a long-duration cine CT for respiratory gated PET/CT, whose cine acquisition time is ten times longer than a traditional short-duration cine CT. We compare the proposed long-duration cine CT with the traditional short-duration cine CT through numerous phantom simulations with 11 respiratory traces measured during patient PET/CT scans. Experimental results show that, the long-duration cine CT reduces the motion mismatch between PET and CT by 41% and improves the overall reconstruction accuracy by 42% on average, as compared to the traditional short-duration cine CT. The long-duration cine CT also reduces artifacts in PET images caused by misalignment and mismatch between adjacent slices in phase-gated CT images. The improvement in motion matching between PET and CT by extending the cine duration depends on the patient, with potentially greater benefits for patients with irregular breathing patterns or larger diaphragm movements.

  20. First Experience With Real-Time EPID-Based Delivery Verification During IMRT and VMAT Sessions

    International Nuclear Information System (INIS)

    Woodruff, Henry C.; Fuangrod, Todsaporn; Van Uytven, Eric; McCurdy, Boyd M.C.; Beek, Timothy van; Bhatia, Shashank; Greer, Peter B.

    2015-01-01

    Purpose: Gantry-mounted megavoltage electronic portal imaging devices (EPIDs) have become ubiquitous on linear accelerators. WatchDog is a novel application of EPIDs, in which the image frames acquired during treatment are used to monitor treatment delivery in real time. We report on the preliminary use of WatchDog in a prospective study of cancer patients undergoing intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) and identify the challenges of clinical adoption. Methods and Materials: At the time of submission, 28 cancer patients (head and neck, pelvis, and prostate) undergoing fractionated external beam radiation therapy (24 IMRT, 4 VMAT) had ≥1 treatment fraction verified in real time (131 fractions or 881 fields). EPID images acquired continuously during treatment were synchronized and compared with model-generated transit EPID images within a frame time (∼0.1 s). A χ comparison was performed to cumulative frames to gauge the overall delivery quality, and the resulting pass rates were reported graphically during treatment delivery. Every frame acquired (500-1500 per fraction) was saved for postprocessing and analysis. Results: The system reported the mean ± standard deviation in real time χ 91.1% ± 11.5% (83.6% ± 13.2%) for cumulative frame χ analysis with 4%, 4 mm (3%, 3 mm) criteria, global over the integrated image. Conclusions: A real-time EPID-based radiation delivery verification system for IMRT and VMAT has been demonstrated that aims to prevent major mistreatments in radiation therapy.

  1. First Experience With Real-Time EPID-Based Delivery Verification During IMRT and VMAT Sessions

    Energy Technology Data Exchange (ETDEWEB)

    Woodruff, Henry C., E-mail: henry.woodruff@newcastle.edu.au [Faculty of Science and Information Technology, School of Mathematical and Physical Sciences, University of Newcastle, New South Wales (Australia); Fuangrod, Todsaporn [Faculty of Engineering and Built Environment, School of Electrical Engineering and Computer Science, University of Newcastle, New South Wales (Australia); Van Uytven, Eric; McCurdy, Boyd M.C.; Beek, Timothy van [Division of Medical Physics, CancerCare Manitoba, Winnipeg, Manitoba (Canada); Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba (Canada); Department of Radiology, University of Manitoba, Winnipeg, Manitoba (Canada); Bhatia, Shashank [Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, New South Wales (Australia); Greer, Peter B. [Faculty of Science and Information Technology, School of Mathematical and Physical Sciences, University of Newcastle, New South Wales (Australia); Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, New South Wales (Australia)

    2015-11-01

    Purpose: Gantry-mounted megavoltage electronic portal imaging devices (EPIDs) have become ubiquitous on linear accelerators. WatchDog is a novel application of EPIDs, in which the image frames acquired during treatment are used to monitor treatment delivery in real time. We report on the preliminary use of WatchDog in a prospective study of cancer patients undergoing intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) and identify the challenges of clinical adoption. Methods and Materials: At the time of submission, 28 cancer patients (head and neck, pelvis, and prostate) undergoing fractionated external beam radiation therapy (24 IMRT, 4 VMAT) had ≥1 treatment fraction verified in real time (131 fractions or 881 fields). EPID images acquired continuously during treatment were synchronized and compared with model-generated transit EPID images within a frame time (∼0.1 s). A χ comparison was performed to cumulative frames to gauge the overall delivery quality, and the resulting pass rates were reported graphically during treatment delivery. Every frame acquired (500-1500 per fraction) was saved for postprocessing and analysis. Results: The system reported the mean ± standard deviation in real time χ 91.1% ± 11.5% (83.6% ± 13.2%) for cumulative frame χ analysis with 4%, 4 mm (3%, 3 mm) criteria, global over the integrated image. Conclusions: A real-time EPID-based radiation delivery verification system for IMRT and VMAT has been demonstrated that aims to prevent major mistreatments in radiation therapy.

  2. Cine y literatura

    Directory of Open Access Journals (Sweden)

    Jorge Enrique Adoum

    2015-01-01

    Full Text Available El lanzamiento de la undécima edición de la novela de Jorge Enrique Adoum "Entre Marx y una mujer desnuda" y el comienzo simultáneo de la filmación de la película basada en esa obra, ha puesto nuevamente de relieve, el recurrente tema de la literatura llevada al cine. En ese sentido, en agosto, se llevó a cabo una mesa redonda en la que participaron como exponentes Camilo Luzuriaga (director del filme, Omar Ospina (crítico cinematográfico y jorge Enrique Adoum. En el siguiente artículo -basado en su exposición en la mesa redonda-, el reconocido escritor analiza cómo debe ser para él la vinculación entre literatura y cine.

  3. Cine e identidades virtuales

    Directory of Open Access Journals (Sweden)

    Labrador Ben, Julia María

    2006-12-01

    Full Text Available The subject is the ability the cinema has performed - since its very begining - as a creation instrument of national and transnational identities. In the first place, it covers Spanish cinema since 1938 and right after the end of the Spanish Civil War, II World War Italian-German cinema, Eisenstein’s, Leni Riefenstahl’s, Griffth’s and Ford’s cinema, the role American war and science fiction productions played, as well as the comedies dealing with the divided Berlin subject. The role of computers and video-games and the new cinema positionings that forecast the end of what is real, is studied through films such as Tron, The Lawnmower Man, Johnny Mnemonic, Ghost in the shell, Simone o Matrix.Se estudia la capacidad del cine desde sus orígenes como instrumento de generación de identidades nacionales y transnacionales. Se realiza, en primer lugar, un recorrido por el cine español desde 1938 y tras el final de la guerra civil, el cine italo-alemán de la II Guerra Mundial, el cine de Eisenstein, Leni Riefenstahl, Griffth y Ford, y el papel que jugaron las producciones norteamericanas bélicas y de ciencia ficción, así como las comedias sobre el Berlín dividido. El papel de los ordenadores y los videojuegos, y los nuevos planteamientos cinematográficos que pronostican el final de lo real se estudia a través de películas como Tron, El cortador de césped, Johnny Mnemonic, Ghost in the shell, Simone o Matrix.

  4. SU-E-T-05: A 2D EPID Transit Dosimetry Model Based On An Empirical Quadratic Formalism

    International Nuclear Information System (INIS)

    Tan, Y; Metwaly, M; Glegg, M; Baggarley, S; Elliott, A

    2014-01-01

    Purpose: To describe a 2D electronic portal imaging device (EPID) transit dosimetry model, based on an empirical quadratic formalism, that can predict either EPID or in-phantom dose distribution for comparisons with EPID captured image or treatment planning system (TPS) dose respectively. Methods: A quadratic equation can be used to relate the reduction in intensity of an exit beam to the equivalent path length of the attenuator. The calibration involved deriving coefficients from a set of dose planes measured for homogeneous phantoms with known thicknesses under reference conditions. In this study, calibration dose planes were measured with EPID and ionisation chamber (IC) in water for the same reference beam (6MV, 100mu, 20×20cm 2 ) and set of thicknesses (0–30cm). Since the same calibration conditions were used, the EPID and IC measurements can be related through the quadratic equation. Consequently, EPID transit dose can be predicted from TPS exported dose planes and in-phantom dose can be predicted using EPID distribution captured during treatment as an input. The model was tested with 4 open fields, 6 wedge fields, and 7 IMRT fields on homogeneous and heterogeneous phantoms. Comparisons were done using 2D absolute gamma (3%/3mm) and results were validated against measurements with a commercial 2D array device. Results: The gamma pass rates for comparisons between EPID measured and predicted ranged from 93.6% to 100.0% for all fields and phantoms tested. Results from this study agreed with 2D array measurements to within 3.1%. Meanwhile, comparisons in-phantom between TPS computed and predicted ranged from 91.6% to 100.0%. Validation with 2D array device was not possible for inphantom comparisons. Conclusion: A 2D EPID transit dosimetry model for treatment verification was described and proven to be accurate. The model has the advantage of being generic and allows comparisons at the EPID plane as well as multiple planes in-phantom

  5. Simultaneous static and cine nonenhanced MR angiography using radial sampling and highly constrained back projection reconstruction.

    Science.gov (United States)

    Koktzoglou, Ioannis; Mistretta, Charles A; Giri, Shivraman; Dunkle, Eugene E; Amin, Parag; Edelman, Robert R

    2014-10-01

    To describe a pulse sequence for simultaneous static and cine nonenhanced magnetic resonance angiography (NEMRA) of the peripheral arteries. The peripheral arteries of 10 volunteers and 6 patients with peripheral arterial disease (PAD) were imaged with the proposed cine NEMRA sequence on a 1.5 Tesla (T) system. The impact of multi-shot imaging and highly constrained back projection (HYPR) reconstruction was examined. The propagation rate of signal along the length of the arterial tree in the cine nonenhanced MR angiograms was quantified. The cine NEMRA sequence simultaneously provided a static MR angiogram showing vascular anatomy as well as a cine display of arterial pulse wave propagation along the entire length of the peripheral arteries. Multi-shot cine NEMRA improved temporal resolution and reduced image artifacts. HYPR reconstruction improved image quality when temporal reconstruction footprints shorter than 100 ms were used (P cine NEMRA was slower in patients with PAD than in volunteers. Simultaneous static and cine NEMRA of the peripheral arteries is feasible. Multi-shot acquisition and HYPR reconstruction can be used to improve arterial conspicuity and temporal resolution. Copyright © 2013 Wiley Periodicals, Inc.

  6. Cine computed tomography for diagnosis of superior vena cava obstruction following the mustard operation

    International Nuclear Information System (INIS)

    Matherne, G.P.; Atkins, D.L.; Frey, E.E.; Smith, W.L.

    1987-01-01

    Superior vena caval obstruction is a well described complication following Mustard's repair for transposition of the great arteries. We report a case of a 6-year-old child with superior vena cava obstruction correctly diagnosed by Cine-CT. The advantages of imaging with Cine-CT for this complication are discussed. (orig.)

  7. Free-breathing steady-state free precession cine cardiac magnetic resonance with respiratory navigator gating.

    Science.gov (United States)

    Moghari, Mehdi H; Komarlu, Rukmini; Annese, David; Geva, Tal; Powell, Andrew J

    2015-04-01

    To develop and validate a respiratory motion compensation method for free-breathing cardiac cine imaging. A free-breathing navigator-gated cine steady-state free precession acquisition (Cine-Nav) was developed which preserves the equilibrium state of the net magnetization vector, maintains the high spatial and temporal resolutions of standard breath-hold (BH) acquisition, and images entire cardiac cycle. Cine image data is accepted only from cardiac cycles occurring entirely during end-expiration. Prospective validation was performed in 10 patients by obtaining in each three complete ventricular image stacks with different respiratory motion compensation approaches: (1) BH, (2) free-breathing with 3 signal averages (3AVG), and (3) free-breathing with Cine-Nav. The subjective image quality score (1 = worst, 4 = best) for Cine-Nav (3.8 ± 0.4) was significantly better than for 3AVG (2.2 ± 0.5, P = 0.002), and similar to BH (4.0 ± 0.0, P = 0.13). The blood-to-myocardium contrast ratio for Cine-Nav (6.3 ± 1.5) was similar to BH (5.9 ± 1.6, P = 0.52) and to 3AVG (5.6 ± 2.5, P = 0.43). There were no significant differences between Cine-Nav and BH for the ventricular volumes and mass. In contrast, there were significant differences between 3AVG and BH in all of these measurements but right ventricular mass. Free-breathing cine imaging with Cine-Nav yielded comparable image quality and ventricular measurements to BH, and was superior to 3AVG. © 2014 Wiley Periodicals, Inc.

  8. Cine MRI: a new approach to the diagnosis of scapholunate dissociation

    Energy Technology Data Exchange (ETDEWEB)

    Langner, I.; Eisenschenk, A. [University Medicine Greifswald, Division of Hand Surgery and Functional Microsurgery, Department of Trauma and Reconstructive Surgery, Greifswald (Germany); Fischer, S.; Langner, S. [University Medicine Greifswald, Institute for Diagnostic Radiology and Neuroradiology, Greifswald (Germany)

    2015-08-15

    To evaluate the feasibility of cine MRI for the detection of scapholunate dissociation (SLD) and to compare the sensitivity and specificity of cine MRI with those of cineradiography and arthroscopy. To evaluate feasibility, healthy subjects underwent cine MRI of the wrist. To evaluate sensitivity and specificity, patients with clinically suspected scapholunate ligament (SLL) injury after trauma to the wrist were prospectively included and underwent radiographic examination, cineradiography, and cine MRI. In 25 out of 38 patients, subsequent arthroscopy was performed. Results of cineradiography and cine MRI correlated with those of arthroscopy. Cine MRI was of diagnostic quality in all healthy subjects and patients with good interrater agreement. There was excellent correlation between cineradiography and cine MRI. Scapholunate distance differed significantly between healthy subjects and patients with scapholunate dissociation (p < 0.001), but not between imaging modalities in the patient group. Cine MRI had 85 % sensitivity and 90 % specificity for the detection of SLD. Cine MRI of the wrist is a fast and reliable technique for the detection of SLD with diagnostic accuracy comparable to cineradiography. It can be easily implemented as a routine clinical MRI examination, facilitating diagnostic workup of patients with suspected SLD while avoiding radiation exposure. (orig.)

  9. The Cine Club de Lima

    Directory of Open Access Journals (Sweden)

    Ricardo Bedoya

    2009-10-01

    Full Text Available Cine Club de Lima was the most important cinematography culture spreading institution of the fifties. It was also the precursor of other movie clubs of following decades, which would bring together important Peruvian intellectuals from every art form, who would together promote cultural entertainment. This piece describes ”Cine Club Lima” origins and dissolution, while taking into account its programs.

  10. Virtual EPID standard phantom audit (VESPA) for remote IMRT and VMAT credentialing

    Science.gov (United States)

    Miri, Narges; Lehmann, Joerg; Legge, Kimberley; Vial, Philip; Greer, Peter B.

    2017-06-01

    A virtual EPID standard phantom audit (VESPA) has been implemented for remote auditing in support of facility credentialing for clinical trials using IMRT and VMAT. VESPA is based on published methods and a clinically established IMRT QA procedure, here extended to multi-vendor equipment. Facilities are provided with comprehensive instructions and CT datasets to create treatment plans. They deliver the treatment directly to their EPID without any phantom or couch in the beam. In addition, they deliver a set of simple calibration fields per instructions. Collected EPID images are uploaded electronically. In the analysis, the dose is projected back into a virtual cylindrical phantom. 3D gamma analysis is performed. 2D dose planes and linear dose profiles are provided and can be considered when needed for clarification. In addition, using a virtual flat-phantom, 2D field-by-field or arc-by-arc gamma analyses are performed. Pilot facilities covering a range of planning and delivery systems have performed data acquisition and upload successfully. Advantages of VESPA are (1) fast turnaround mainly driven by the facility’s capability of providing the requested EPID images, (2) the possibility for facilities performing the audit in parallel, as there is no need to wait for a phantom, (3) simple and efficient credentialing for international facilities, (4) a large set of data points, and (5) a reduced impact on resources and environment as there is no need to transport heavy phantoms or audit staff. Limitations of the current implementation of VESPA for trials credentialing are that it does not provide absolute dosimetry, therefore a Level I audit is still required, and that it relies on correctly delivered open calibration fields, which are used for system calibration. The implemented EPID based IMRT and VMAT audit system promises to dramatically improve credentialing efficiency for clinical trials and wider applications.

  11. Cine MRI of patients with cervical myelopathy

    International Nuclear Information System (INIS)

    Ukita, Yasutaka

    1993-01-01

    Forty-six patients with cervical myelopathy were examined before and after surgery by cine magnetic resonance imaging (MRI). According to the occurrence site and degree of flow void, cerebrospinal fluid (CSF) flow void was classified into five: anterior type (flow void mainly in the anterior part of subarachnoid space), posterior type (mainly in the posteiror part), anteroposterior type (in the anterior and posterior parts), incomplete block type (flow void limited to the upper and lower parts of the block), and complete block type (no flow void). None of the 46 patients had normal CSF flow void on cine MRI before surgery. CSF flow void was seen in systolic phase on ECG (from 150 to 300 msec from R's wave) in all patients after spinal cord decompression. Postoperative CBF flow void types correlated well with surgical method, disease, and postoperative vertebral alignment. Postoperative outcome was the most excellent in the group of posterior type and the poorest in the group of anteroposterior type, showing a significant difference between the groups. Cine MRI is a useful noninvasive, dynamic method for assessing postoperative decompression effect. (N.K.)

  12. CineGlob

    CERN Multimedia

    CineGlob

    2014-01-01

    CERN will be hosting the next CineGlobe International Film Festival from March 18 to 23 at the Globe of Science and Innovation with the theme “Beyond the Frontier”, This 4th edition will see 66 short films “inspired by science” in competition, including fiction films as well as documentaries. From Tuesday the 18th to Friday the 21st, special lunch sessions are organized from 12:30 to 13:30. Food will be available for purchase at the tent “Café Cinéma” next to the Globe. Special evenings While the short film screenings are the heart of the festival, every year CineGlobe also organizes compelling special events every night, from feature films to special musical performances: Tuesday 18th: The Swiss premiere of the acclaimed documentary film Particle Fever. Presented by BBC Storyville with Fabiola Gianotti and director Mark Levinson in attendance; Wednesday 19th: The result of the “Story Matter” Hacka...

  13. Kinematic analysis of the knee joint by cine MRI

    International Nuclear Information System (INIS)

    Niitsu, Mamoru; Akisada, Masayoshi; Anno, Izumi; Matsumoto, Kunihiko; Kuno, Shin-ya; Miyakawa, Shunpei; Inouye, Tamon; Kose, Katsumi.

    1989-01-01

    In order to obtain the MR imaging of a moving knee joint, we developed a drive system of the knee. A reciprocating reversible motor with a rope and pulleys drove a knee brace with the knee bending and extending every two seconds. Using photo sensor probe for gating cine acquisition, we got 16-time frames/cycle MR images. Such as articular cartilage, ligaments and synovial fluid, the fine components of a moving knee joint were clearly seen. In a dynamic display, these cine images demonstrated 'actual' movement of the knee joint. Moving joint fluid and defect of anterior cruciate ligament were demonstrated in the case of knee injury. These findings were not seen on static images. Cine MR imaging was also helpful for evaluating the chronic joint disease and ligament reconstruction. Through the use of the present drive system and cine acquisition, dynamic MR imaging of a moving knee joint is clearly demonstrated and it may provide useful information in the kinematic analysis of the normal and pathologic knee. (author)

  14. Intramyocardial strain estimation from cardiac cine MRI.

    Science.gov (United States)

    Elnakib, Ahmed; Beache, Garth M; Gimel'farb, Georgy; El-Baz, Ayman

    2015-08-01

    Functional strain is one of the important clinical indicators for the quantification of heart performance and the early detection of cardiovascular diseases, and functional strain parameters are used to aid therapeutic decisions and follow-up evaluations after cardiac surgery. A comprehensive framework for deriving functional strain parameters at the endocardium, epicardium, and mid-wall of the left ventricle (LV) from conventional cine MRI data was developed and tested. Cine data were collected using short TR-/TE-balanced steady-state free precession acquisitions on a 1.5T Siemens Espree scanner. The LV wall borders are segmented using a level set-based deformable model guided by a stochastic force derived from a second-order Markov-Gibbs random field model that accounts for the object shape and appearance features. Then, the mid-wall of the segmented LV is determined based on estimating the centerline between the endocardium and epicardium of the LV. Finally, a geometrical Laplace-based method is proposed to track corresponding points on successive myocardial contours throughout the cardiac cycle in order to characterize the strain evolutions. The method was tested using simulated phantom images with predefined point locations of the LV wall throughout the cardiac cycle. The method was tested on 30 in vivo datasets to evaluate the feasibility of the proposed framework to index functional strain parameters. The cine MRI-based model agreed with the ground truth for functional metrics to within 0.30 % for indexing the peak systolic strain change and 0.29 % (per unit time) for indexing systolic and diastolic strain rates. The method was feasible for in vivo extraction of functional strain parameters. Strain indexes of the endocardium, mid-wall, and epicardium can be derived from routine cine images using automated techniques, thereby improving the utility of cine MRI data for characterization of myocardial function. Unlike traditional texture-based tracking, the

  15. Preliminary studies for implementation of a MCL quality control using EPID (Portal Dosimetry); Estudos preliminares para implementacao de um controle de qualidade de MLC com o uso do EPID (Portal Dosimetry)

    Energy Technology Data Exchange (ETDEWEB)

    Mattos, Fabio R.; Furnari, Laura [Universidade de Sao Paulo (USP), Sao Paulo, SP (Brazil). Faculdade de Medicina

    2016-07-01

    A Quality Control (CQ) to ensure the expected performance of a Multileaf Collimator System (MLC) is essential for delivering dose in a safety and appropriate way. The time required for equipment control and dosimetry may be lowered when the Electronic Portal Image Device (EPID) is used. The aim of this paper was to check the resolution limits of the detection system for IMRT mode, and to do the analysis of three tests of MLC performance: Picket Fence, Slinding GAP, MLC versus Gantry. A Varian iX Clinac equipped with an 80 leaf Millennium MLC and with amorphous silicon based EPID (aS1000) was use. The EPID proved Effective, where errors up to 0,5 mm can be detected. Information about interleaf transmissions, dose profile and gravity influence in the leaf banks also were included. (author)

  16. A comprehensive study of the mechanical performance of gantry, EPID and the MLC assembly in Elekta linacs during gantry rotation

    DEFF Research Database (Denmark)

    Rowshanfarzad, P; Lynggaard Riis, Hans; Zimmermann, S J

    2015-01-01

    OBJECTIVE: In radiotherapy treatments, it is crucial to monitor the performance of linear accelerator (linac) components, including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested...... collimator leaf bank assemblies was around 1 mm. A meaningful correlation was found between the age of the linacs and their mechanical performance. Conclusions and Advances in knowledge: The method and software developed in this study provide a simple tool for effective investigation of the behaviour...

  17. Robustness and precision of an automatic marker detection algorithm for online prostate daily targeting using a standard V-EPID.

    Science.gov (United States)

    Aubin, S; Beaulieu, L; Pouliot, S; Pouliot, J; Roy, R; Girouard, L M; Martel-Brisson, N; Vigneault, E; Laverdière, J

    2003-07-01

    An algorithm for the daily localization of the prostate using implanted markers and a standard video-based electronic portal imaging device (V-EPID) has been tested. Prior to planning, three gold markers were implanted in the prostate of seven patients. The clinical images were acquired with a BeamViewPlus 2.1 V-EPID for each field during the normal course radiotherapy treatment and are used off-line to determine the ability of the automatic marker detection algorithm to adequately and consistently detect the markers. Clinical images were obtained with various dose levels from ranging 2.5 to 75 MU. The algorithm is based on marker attenuation characterization in the portal image and spatial distribution. A total of 1182 clinical images were taken. The results show an average efficiency of 93% for the markers detected individually and 85% for the group of markers. This algorithm accomplishes the detection and validation in 0.20-0.40 s. When the center of mass of the group of implanted markers is used, then all displacements can be corrected to within 1.0 mm in 84% of the cases and within 1.5 mm in 97% of cases. The standard video-based EPID tested provides excellent marker detection capability even with low dose levels. The V-EPID can be used successfully with radiopaque markers and the automatic detection algorithm to track and correct the daily setup deviations due to organ motions.

  18. Quality control program of multi-leaf collimation based EPID for teams with Rapidarc

    International Nuclear Information System (INIS)

    Pujades Claumarchirant, M. C.; Richart Sancho, J.; Gimeno Olmos, J.; Lliso Valverde, F.; Carmona Mesenguer, V.; Garcia Martinez, M. T.; Palomo Llinares, R.; Ballester Pallares, F.; Perez Calatayud, J.

    2013-01-01

    The objective of this work is to show a collection of different recommendations on the control of quality of collimation multi-leaf system and present the selection of tests based on the electronic imaging device (EPID) portal that have decided to establish in our Center, where in addition to the requirements of quality assurance generic for collimation multi-leaf system quality control methods have been included for RapidArc. (Author)

  19. Comparison Between the Spectrum reconstruction methodology and release transmitted by both based in pictures taken EPID; Comparacion entre la metodologia de recostruccion de espectros por transmision y por disperison, ambas basadas en imagenes tomadas con EPID

    Energy Technology Data Exchange (ETDEWEB)

    Juste, B.; Miro, R.; Jambrina, A.; Campayo, J. M.; Diez, S.; Santos, A.; Verdu, V.

    2013-07-01

    A comparison of the spectral reconstruction based on data transmission and spectral reconstruction based on scattering data is presented, we have both developed using EPID images. It is shown that the reconstruction results based on transmission offer much better fit with the theoretical predictions.

  20. MO-AB-BRA-03: Development of Novel Real Time in Vivo EPID Treatment Verification for Brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Fonseca, G; Podesta, M [Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht 6201 BN (Netherlands); Reniers, B [Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht 6201 BN (Netherlands); Research Group NuTeC, CMK, Hasselt University, Agoralaan Gebouw H, Diepenbeek B-3590 (Belgium); Verhaegen, F [Department of Radiation Oncology (MAASTRO), GROW School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht 6201 BN (Netherlands); Medical Physics Unit, Department of Oncology, McGill University, Montreal, Quebec H3G 1A4 (Canada)

    2016-06-15

    Purpose: High Dose Rate (HDR) brachytherapy treatments are employed worldwide to treat a wide variety of cancers. However, in vivo dose verification remains a challenge with no commercial dosimetry system available to verify the treatment dose delivered to the patient. We propose a novel dosimetry system that couples an independent Monte Carlo (MC) simulation platform and an amorphous silicon Electronic Portal Imaging Device (EPID) to provide real time treatment verification. Methods: MC calculations predict the EPID response to the photon fluence emitted by the HDR source by simulating the patient, the source dwell positions and times, and treatment complexities such as tissue compositions/densities and different applicators. Simulated results are then compared against EPID measurements acquired with ∼0.14s time resolution which allows dose measurements for each dwell position. The EPID has been calibrated using an Ir-192 HDR source and experiments were performed using different phantoms, including tissue equivalent materials (PMMA, lung and bone). A source positioning accuracy of 0.2 mm, without including the afterloader uncertainty, was ensured using a robotic arm moving the source. Results: An EPID can acquire 3D Cartesian source positions and its response varies significantly due to differences in the material composition/density of the irradiated object, allowing detection of changes in patient geometry. The panel time resolution allows dose rate and dwell time measurements. Moreover, predicted EPID images obtained from clinical treatment plans provide anatomical information that can be related to the patient anatomy, mostly bone and air cavities, localizing the source inside of the patient using its anatomy as reference. Conclusion: Results obtained show the feasibility of the proposed dose verification system that is capable to verify all the brachytherapy treatment steps in real time providing data about treatment delivery quality and also applicator

  1. Development and implementation of EPID-based quality assurance tests for the small animal radiation research platform (SARRP).

    Science.gov (United States)

    Anvari, Akbar; Poirier, Yannick; Sawant, Amit

    2018-04-28

    Although small animal image-guided radiotherapy (SA-IGRT) systems are used increasingly in preclinical research, tools for performing routine quality assurance (QA) have not been optimized and are not readily available. Robust, efficient, and reliable QA tools are needed to ensure the accuracy and reproducibility of SA-IGRT systems. Several investigators have reported custom-made phantoms and protocols for SA-IGRT systems QA. These are typically time and resource intensive and are therefore not well suited to the preclinical radiotherapy environment, in which physics support is limited and routine QA is performed by technical staff. We investigated the use of the inbuilt electronic portal imaging device (EPID) to develop and validate routine QA tests and procedures. In this work, we focus on the Xstrahl Small Animal Radiation Research Platform (SARRP) EPID. However, the methodology and tests developed here are applicable to any SA-IGRT system that incorporates an EPID. We performed a comprehensive characterization of the dosimetric properties of the camera-based EPID at kilovoltage energies over a 11-month period, including detector warm-up time, radiation dose history effect, stability and short- and long-term reproducibility, gantry angle dependency, output factor, and linearity of the EPID response. We developed a test to measure the constancy of beam quality in terms of half-value layer and tube peak potential using the EPID. We verified the SARRP daily output and beam profile constancy using the imager. We investigated the use of the imager to monitor beam-targeting accuracy at various gantry and couch angles. The EPID response was stable and reproducible, exhibiting maximum variations of ≤0.3% and ≤1.9% for short and long terms, respectively. The detector showed no dependence on response at different gantry angles, with a maximum variation ≤0.5%. We found close agreement in output factor measurement between the portal imager and reference dosimeters

  2. Cine MRI of swallowing in patients with advanced oral or oropharyngeal carcinoma: a feasibility study

    NARCIS (Netherlands)

    Kreeft, Anne Marijn; Rasch, Coen R. N.; Muller, Sara H.; Pameijer, Frank A.; Hallo, Eeke; Balm, Alfons J. M.

    2012-01-01

    Treatment of oral and oropharyngeal cancer may cause dysphagia. Purpose is to examine whether cine magnetic resonance imaging (MRI) yields additional information compared to standard examination in the evaluation of posttreatment dysphagia and mobility of oral and oropharyngeal structures.

  3. Cine dyscontractility index: A novel marker of mechanical dyssynchrony that predicts response to cardiac resynchronization therapy.

    Science.gov (United States)

    Werys, Konrad; Petryka-Mazurkiewicz, Joanna; Błaszczyk, Łukasz; Miśko, Jolanta; Śpiewak, Mateusz; Małek, Łukasz A; Mazurkiewicz, Łukasz; Miłosz-Wieczorek, Barbara; Marczak, Magdalena; Kubik, Agata; Dąbrowska, Agnieszka; Piątkowska-Janko, Ewa; Sawionek, Błażej; Wijesurendra, Rohan; Piechnik, Stefan K; Bogorodzki, Piotr

    2016-12-01

    To investigate whether magnetic resonance imaging (MRI) cine-derived dyssynchrony indices provide additional information compared to conventional tagged MRI (tMRI) acquisitions in heart failure patients undergoing cardiac resynchronization therapy (CRT). Patients scheduled for CRT (n = 52) underwent preprocedure MRI including cine and tMRI acquisitions. Segmental strain curves were calculated for both cine and tMRI to produce a range of standard indices for direct comparison between modalities. We also proposed and evaluated a novel index of "dyscontractility," which detects the presence of focal areas with paradoxically positive circumferential strain. Across conventional strain indices, there was only moderate-to-poor (R = 0.3-0.6) correlation between modalities; eight cine-derived indices showed statistically significant (P cine images (cine dyscontractility index, "CDI") was the single best predictor of clinical response to CRT (area under the curve AUC = 0.81, P Cine-derived strain indices offer potentially new information compared to tMRI. Specifically, the novel CDI is most strongly linked to response to cardiac resynchronization therapy in a contemporary patient cohort. It utilizes readily available MRI data, is relatively straightforward to process, and compares favorably with any conventional tagging index. J. Magn. Reson. Imaging 2016;44:1483-1492. © 2016 International Society for Magnetic Resonance in Medicine.

  4. Cine-Club

    CERN Multimedia

    Cine-Club

    2011-01-01

    Thursday 10 November 2011 at 20:30 CERN Council Chamber The Counterfeiters / Die Fälscher By: Stefan Ruzowitzky (Germany/Austria, 2007) 98 min With: Karl Markovics, August Diel, Devid Striesow The Counterfeiters is the true story of the largest counterfeiting operation in history, set up by the Nazis in 1936. Master counterfeiter Salomon ‘Sally’ Sorowitsch, a Jew without morals, is arrested and sent to Mauthausen concentration camp, where he weasels through by offering his ‘artistic’ services. After transfer to Sachsenhausen he’s given an official alternative: taking ‘artistic’ charge of a cons teams assigned to produce perfect forgeries of allied banknotes for the Third Reich. Survivor Sally accepts, but conscience matters among the team prove no less dangerous then a sadistic jailer, which only the CO may keep in check. Original version German; English subtitles Entrance: 2 CHF Projection from DVD http://cine...

  5. CINE: Comet INfrared Excitation

    Science.gov (United States)

    de Val-Borro, Miguel; Cordiner, Martin A.; Milam, Stefanie N.; Charnley, Steven B.

    2017-08-01

    CINE calculates infrared pumping efficiencies that can be applied to the most common molecules found in cometary comae such as water, hydrogen cyanide or methanol. One of the main mechanisms for molecular excitation in comets is the fluorescence by the solar radiation followed by radiative decay to the ground vibrational state. This command-line tool calculates the effective pumping rates for rotational levels in the ground vibrational state scaled by the heliocentric distance of the comet. Fluorescence coefficients are useful for modeling rotational emission lines observed in cometary spectra at sub-millimeter wavelengths. Combined with computational methods to solve the radiative transfer equations based, e.g., on the Monte Carlo algorithm, this model can retrieve production rates and rotational temperatures from the observed emission spectrum.

  6. Cine Letras Goytisolo Cartelera

    Directory of Open Access Journals (Sweden)

    Flávio Pimenta de Souza

    2011-10-01

    Full Text Available Resumo: Estudo de investigação sobre a obra literária do escritor Juan Goytisolo em que se analisam comparativamente as técnicas cinematográficas do projeto Dogma 95 e as técnicas literárias do referido autor.Palavras-chave: Literatura espanhola; Juan Goytisolo; cinema.Resumen: Estúdio de investigación sobre la obra literaria del escritor Juan Goytisolo en el que se analizan comparativamente las técnicas cinematográficas del proyecto Dogma 95 y las técnicas literarias de dicho autor.Palabras-clave: Literatura española; Juan Goytisolo; cine.Keywords: Spanish literature; Juan Goytisolo; cinema.

  7. Registration of clinical volumes to beams-eye-view images for real-time tracking

    Energy Technology Data Exchange (ETDEWEB)

    Bryant, Jonathan H.; Rottmann, Joerg; Lewis, John H.; Mishra, Pankaj; Berbeco, Ross I., E-mail: rberbeco@lroc.harvard.edu [Department of Radiation Oncology, Brigham and Women’s Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, Massachusetts 02115 (United States); Keall, Paul J. [Radiation Physics Laboratory, Sydney Medical School, University of Sydney, Sydney, New South Wales 2006 (Australia)

    2014-12-15

    Purpose: The authors combine the registration of 2D beam’s eye view (BEV) images and 3D planning computed tomography (CT) images, with relative, markerless tumor tracking to provide automatic absolute tracking of physician defined volumes such as the gross tumor volume (GTV). Methods: During treatment of lung SBRT cases, BEV images were continuously acquired with an electronic portal imaging device (EPID) operating in cine mode. For absolute registration of physician-defined volumes, an intensity based 2D/3D registration to the planning CT was performed using the end-of-exhale (EoE) phase of the four dimensional computed tomography (4DCT). The volume was converted from Hounsfield units into electron density by a calibration curve and digitally reconstructed radiographs (DRRs) were generated for each beam geometry. Using normalized cross correlation between the DRR and an EoE BEV image, the best in-plane rigid transformation was found. The transformation was applied to physician-defined contours in the planning CT, mapping them into the EPID image domain. A robust multiregion method of relative markerless lung tumor tracking quantified deviations from the EoE position. Results: The success of 2D/3D registration was demonstrated at the EoE breathing phase. By registering at this phase and then employing a separate technique for relative tracking, the authors are able to successfully track target volumes in the BEV images throughout the entire treatment delivery. Conclusions: Through the combination of EPID/4DCT registration and relative tracking, a necessary step toward the clinical implementation of BEV tracking has been completed. The knowledge of tumor volumes relative to the treatment field is important for future applications like real-time motion management, adaptive radiotherapy, and delivered dose calculations.

  8. An EPID response calculation algorithm using spatial beam characteristics of primary, head scattered and MLC transmitted radiation

    International Nuclear Information System (INIS)

    Rosca, Florin; Zygmanski, Piotr

    2008-01-01

    We have developed an independent algorithm for the prediction of electronic portal imaging device (EPID) response. The algorithm uses a set of images [open beam, closed multileaf collimator (MLC), various fence and modified sweeping gap patterns] to separately characterize the primary and head-scatter contributions to EPID response. It also characterizes the relevant dosimetric properties of the MLC: Transmission, dosimetric gap, MLC scatter [P. Zygmansky et al., J. Appl. Clin. Med. Phys. 8(4) (2007)], inter-leaf leakage, and tongue and groove [F. Lorenz et al., Phys. Med. Biol. 52, 5985-5999 (2007)]. The primary radiation is modeled with a single Gaussian distribution defined at the target position, while the head-scatter radiation is modeled with a triple Gaussian distribution defined downstream of the target. The distances between the target and the head-scatter source, jaws, and MLC are model parameters. The scatter associated with the EPID is implicit in the model. Open beam images are predicted to within 1% of the maximum value across the image. Other MLC test patterns and intensity-modulated radiation therapy fluences are predicted to within 1.5% of the maximum value. The presented method was applied to the Varian aS500 EPID but is designed to work with any planar detector with sufficient spatial resolution

  9. A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment.

    Science.gov (United States)

    Fuangrod, Todsaporn; Woodruff, Henry C; van Uytven, Eric; McCurdy, Boyd M C; Kuncic, Zdenka; O'Connor, Daryl J; Greer, Peter B

    2013-09-01

    To design and develop a real-time electronic portal imaging device (EPID)-based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient. The system utilizes a comprehensive physics-based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real-time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame-by-frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance. The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real-time gamma results for frame-by-frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID-based real-time delivery verification system successfully detected simulated gross errors introduced into patient plan deliveries in near real-time (within 0.1 s). A real-time radiation delivery verification system for dynamic IMRT has been demonstrated that is designed to prevent major mistreatments in modern radiation therapy.

  10. A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment

    Energy Technology Data Exchange (ETDEWEB)

    Fuangrod, Todsaporn [Faculty of Engineering and Built Environment, School of Electrical Engineering and Computer Science, the University of Newcastle, NSW 2308 (Australia); Woodruff, Henry C.; O’Connor, Daryl J. [Faculty of Science and IT, School of Mathematical and Physical Sciences, the University of Newcastle, NSW 2308 (Australia); Uytven, Eric van; McCurdy, Boyd M. C. [Division of Medical Physics, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada); Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2 (Canada); Department of Radiology, University of Manitoba, Winnipeg, Manitoba R3T 2N2 (Canada); Kuncic, Zdenka [School of Physics, University of Sydney, Sydney, NSW 2006 (Australia); Greer, Peter B. [Faculty of Science and IT, School of Mathematical and Physical Sciences, the University of Newcastle, NSW 2308, Australia and Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Locked Bag 7, Hunter region Mail Centre, Newcastle, NSW 2310 (Australia)

    2013-09-15

    Purpose: To design and develop a real-time electronic portal imaging device (EPID)-based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient.Methods: The system utilizes a comprehensive physics-based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real-time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame-by-frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance.Results: The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real-time gamma results for frame-by-frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID-based real-time delivery verification system successfully detected simulated gross errors introduced into patient plan deliveries in near real-time (within 0.1 s).Conclusions: A real-time radiation delivery verification system for dynamic IMRT has been demonstrated that is designed to prevent major mistreatments in modern radiation therapy.

  11. A system for EPID-based real-time treatment delivery verification during dynamic IMRT treatment

    International Nuclear Information System (INIS)

    Fuangrod, Todsaporn; Woodruff, Henry C.; O’Connor, Daryl J.; Uytven, Eric van; McCurdy, Boyd M. C.; Kuncic, Zdenka; Greer, Peter B.

    2013-01-01

    Purpose: To design and develop a real-time electronic portal imaging device (EPID)-based delivery verification system for dynamic intensity modulated radiation therapy (IMRT) which enables detection of gross treatment delivery errors before delivery of substantial radiation to the patient.Methods: The system utilizes a comprehensive physics-based model to generate a series of predicted transit EPID image frames as a reference dataset and compares these to measured EPID frames acquired during treatment. The two datasets are using MLC aperture comparison and cumulative signal checking techniques. The system operation in real-time was simulated offline using previously acquired images for 19 IMRT patient deliveries with both frame-by-frame comparison and cumulative frame comparison. Simulated error case studies were used to demonstrate the system sensitivity and performance.Results: The accuracy of the synchronization method was shown to agree within two control points which corresponds to approximately ∼1% of the total MU to be delivered for dynamic IMRT. The system achieved mean real-time gamma results for frame-by-frame analysis of 86.6% and 89.0% for 3%, 3 mm and 4%, 4 mm criteria, respectively, and 97.9% and 98.6% for cumulative gamma analysis. The system can detect a 10% MU error using 3%, 3 mm criteria within approximately 10 s. The EPID-based real-time delivery verification system successfully detected simulated gross errors introduced into patient plan deliveries in near real-time (within 0.1 s).Conclusions: A real-time radiation delivery verification system for dynamic IMRT has been demonstrated that is designed to prevent major mistreatments in modern radiation therapy

  12. Self-gated fat-suppressed cardiac cine MRI.

    Science.gov (United States)

    Ingle, R Reeve; Santos, Juan M; Overall, William R; McConnell, Michael V; Hu, Bob S; Nishimura, Dwight G

    2015-05-01

    To develop a self-gated alternating repetition time balanced steady-state free precession (ATR-SSFP) pulse sequence for fat-suppressed cardiac cine imaging. Cardiac gating is computed retrospectively using acquired magnetic resonance self-gating data, enabling cine imaging without the need for electrocardiogram (ECG) gating. Modification of the slice-select rephasing gradients of an ATR-SSFP sequence enables the acquisition of a one-dimensional self-gating readout during the unused short repetition time (TR). Self-gating readouts are acquired during every TR of segmented, breath-held cardiac scans. A template-matching algorithm is designed to compute cardiac trigger points from the self-gating signals, and these trigger points are used for retrospective cine reconstruction. The proposed approach is compared with ECG-gated ATR-SSFP and balanced steady-state free precession in 10 volunteers and five patients. The difference of ECG and self-gating trigger times has a variability of 13 ± 11 ms (mean ± SD). Qualitative reviewer scoring and ranking indicate no statistically significant differences (P > 0.05) between self-gated and ECG-gated ATR-SSFP images. Quantitative blood-myocardial border sharpness is not significantly different among self-gated ATR-SSFP ( 0.61±0.15 mm -1), ECG-gated ATR-SSFP ( 0.61±0.15 mm -1), or conventional ECG-gated balanced steady-state free precession cine MRI ( 0.59±0.15 mm -1). The proposed self-gated ATR-SSFP sequence enables fat-suppressed cardiac cine imaging at 1.5 T without the need for ECG gating and without decreasing the imaging efficiency of ATR-SSFP. © 2014 Wiley Periodicals, Inc.

  13. Value of cine MRI for better visualization of the proximal small bowel in normal individuals

    International Nuclear Information System (INIS)

    Torkzad, Michael R.; Blomqvist, Lennart; Vargas, Roberto; Tanaka, Chikako

    2007-01-01

    While enteroclysis seems to be the most efficacious method in achieving bowel distension, enterographic methods have become widespread due to the unpleasantness of enteroclysis and the radiation involved with positioning the catheter. Cine images in MRI can be done without radiation. Our aim is to see if and how cine MR imaging can improve visualization of bowel loops by capturing them while distended. Ten healthy individuals were asked to drink up to 2,000 ml of an oral solution made locally over a 60-min period. Then they underwent MRI using coronal balanced fast field echo (b-FFE) covering small bowel loops. If the initial exam revealed collapsed bowel loops an additional 50 mg of erythromycine was given intravenously with the subject still in the scanner and then cine imaging was performed. The degree of distension of different segments of the small bowel was measured before and after cine imaging and compared. The distension score was significantly higher after addition of the cine images as well, being only significant for depiction of the duodenum and jejunum. Our preliminary study suggests that cine MRI can give better image depiction of the proximal small bowel in healthy volunteers, perhaps circumventing the need for enteroclysis in some cases. There is a need for validation of these results in patients with small bowel disease. (orig.)

  14. Electronic portal imaging devices

    International Nuclear Information System (INIS)

    Lief, Eugene

    2008-01-01

    The topics discussed include, among others, the following: Role of portal imaging; Port films vs. EPID; Image guidance: Elekta volume view; Delivery verification; Automation tasks of portal imaging; Types of portal imaging (Fluorescent screen, mirror, and CCD camera-based imaging; Liquid ion chamber imaging; Amorpho-silicon portal imagers; Fluoroscopic portal imaging; Kodak CR reader; and Other types of portal imaging devices); QA of EPID; and Portal dosimetry (P.A.)

  15. Megavoltage conebeam CT cine as final verification of treatment plan in lung stereotactic body radiotherapy

    International Nuclear Information System (INIS)

    Kudithipudi, Vijay; Gayou, Olivier; Colonias, Athanasios

    2016-01-01

    To analyse the clinical impact of megavoltage conebeam computed tomography (MV-CBCT) cine on internal target volume (ITV) coverage in lung stereotactic body radiotherapy (SBRT). One hundred and six patients received lung SBRT. All underwent 4D computed tomography simulation followed by treatment via image guided 3D conformal or intensity modulated radiation. Prior to SBRT, all patients underwent MV-CBCT cine, in which raw projections are displayed as beam's-eye-view fluoroscopic series with the planning target volume (PTV) projected onto each image, enabling verification of tumour motion relative to the PTV and assessment of adequacy of treatment margin. Megavoltage conebeam computed tomography cine was completed 1–2 days prior to SBRT. Four patients (3.8%) had insufficient ITV coverage inferiorly at cine review. All four plans were changed by adding 5 mm on the PTV margin inferiorly. The mean change in PTV volumes was 3.9 cubic centimetres (cc) (range 1.85–6.32 cc). Repeat cine was performed after plan modification to ensure adequate PTV coverage in the modified plans. PTV margin was adequate in the majority of patients with this technique. MV-CBCT cine did show insufficient coverage in a small subset of patients. Insufficient PTV margins may be a function of 4D CT simulation inadequacies or deficiencies in visualizing the ITV inferior border in the full-inhale phase. MV-CBCT cine is a valuable tool for final verification of PTV margins.

  16. Left Ventricle: Fully Automated Segmentation Based on Spatiotemporal Continuity and Myocardium Information in Cine Cardiac Magnetic Resonance Imaging (LV-FAST

    Directory of Open Access Journals (Sweden)

    Lijia Wang

    2015-01-01

    Full Text Available CMR quantification of LV chamber volumes typically and manually defines the basal-most LV, which adds processing time and user-dependence. This study developed an LV segmentation method that is fully automated based on the spatiotemporal continuity of the LV (LV-FAST. An iteratively decreasing threshold region growing approach was used first from the midventricle to the apex, until the LV area and shape discontinued, and then from midventricle to the base, until less than 50% of the myocardium circumference was observable. Region growth was constrained by LV spatiotemporal continuity to improve robustness of apical and basal segmentations. The LV-FAST method was compared with manual tracing on cardiac cine MRI data of 45 consecutive patients. Of the 45 patients, LV-FAST and manual selection identified the same apical slices at both ED and ES and the same basal slices at both ED and ES in 38, 38, 38, and 41 cases, respectively, and their measurements agreed within -1.6±8.7 mL, -1.4±7.8 mL, and 1.0±5.8% for EDV, ESV, and EF, respectively. LV-FAST allowed LV volume-time course quantitatively measured within 3 seconds on a standard desktop computer, which is fast and accurate for processing the cine volumetric cardiac MRI data, and enables LV filling course quantification over the cardiac cycle.

  17. Discursos subversivos en el cine no hablado

    Directory of Open Access Journals (Sweden)

    Rodríguez Fernández, M.ª del Carmen

    2012-12-01

    Full Text Available “Subversive discourses in silent films” aims to analyse the gender transgression that emerges in certain movies of the silent period. These being silent films, their meaning can only be understood from the content of the images. Directors of the silent period were aware of the importance of images and of their power to put across their views to spectators. The field of filmmaking was new but it was not ungendered, and women filmmakers knew that their films were the vehicle to make spectators aware of their feminist positions. Le matelas alcoolique, La souriante Mme. Beudet and Dora Films of America Shorts are examples of this privilege since, although they belong to two periods of silent film history and have different aims, the directors and the films selected are good examples to be revisited with a gender perspective.

    “Discursos subversivos en el cine no hablado” se propone analizar las transgresiones de género que aparecen en algunas películas del cine mudo. Por sus características al tratarse de un cine no hablado, lo que sugieren algunas imágenes es fundamental para entender los significados de las películas. Las directoras de aquella época eran conocedoras de la importancia de las imágenes y del poder que les confería la cámara para expresar sus reivindicaciones al público espectador. El campo de la cinematografía era nuevo, pero tenía género y las directoras sabían que sus películas eran el vehículo para dar a conocer su articulación como feministas. Le matelas alcoolique, La souriante Mme. Beudet y Dora Films of America Shorts son ejemplos de este privilegio puesto que, aunque las películas de esta muestra pertenezcan a dos etapas del cine mudo y de que sus propósitos fueran diferentes, las directoras y las películas seleccionadas son ejemplos merecedores de una revisión con perspectiva de género.

  18. MLC quality assurance using EPID: A fitting technique with subpixel precision

    International Nuclear Information System (INIS)

    Mamalui-Hunter, Maria; Li, Harold; Low, Daniel A.

    2008-01-01

    Amorphous silicon based electronic portal imaging devices (EPIDs) have been shown to be a good alternative to radiographic film for routine quality assurance (QA) of multileaf collimator (MLC) positioning accuracy. In this work, we present a method of acquiring an EPID image of a traditional strip-test image using analytical fits of the interleaf and leaf abutment image signatures. After exposure, the EPID image pixel values are divided by an open field image to remove EPID response and radiation field variations. Profiles acquired in the direction orthogonal to the leaf motion exhibit small peaks caused by interleaf leakage. Gaussian profiles are fitted to the interleaf leakage peaks, the results of which are, using multiobjective optimization, used to calculate the image rotational angle with respect to the collimator axis of rotation. The relative angle is used to rotate the image to align the MLC leaf travel to the image pixel axes. The leaf abutments also present peaks that are fitted by heuristic functions, in this case modified Lorentzian functions. The parameters of the Lorentzian functions are used to parameterize the leaf gap width and positions. By imaging a set of MLC fields with varying gaps forming symmetric and asymmetric abutments, calibration curves with regard to relative peak height (RPH) versus nominal gap width are obtained. Based on this calibration data, the individual leaf positions are calculated to compare with the nominal programmed positions. The results demonstrate that the collimator rotation angle can be determined as accurate as 0.01 deg. . A change in MLC gap width of 0.2 mm leads to a change in RPH of about 10%. For asymmetrically produced gaps, a 0.2 mm MLC leaf gap width change causes 0.2 pixel peak position change. Subpixel resolution is obtained by using a parameterized fit of the relatively large abutment peaks. By contrast, for symmetrical gap changes, the peak position remains unchanged with a standard deviation of 0

  19. Analysis of blowout fractures using cine mode MRI

    International Nuclear Information System (INIS)

    Kawahara, Masaaki; Shiihara, Kumiko; Kimura, Hisashi; Fukai, Sakuko; Tabuchi, Akio; Kojo, Tuyoshi

    1995-01-01

    By observing conventional CT and MRI images, it is difficult to distinguish extension failure from adhesion, bone fracture or damage to the extraocular muscle, any one of which may be the direct cause of the eye movement disturbance accompanying blowout fracture. We therefore carried out dynamic analysis of eye movement disturbance using a cine mode MRI. We put seven fixation points in the gantry of the MRI and filmed eye movement disturbances by the gradient echo method, using a surface coil and holding the vision on each fixation point. We also video recorded the CRT monitor of the MRI to obtain dynamic MRI images. The subjects comprised 5 cases (7-23 years old). In 4 cases, we started orthoptic treatment, saccadic eye movement training, convergence training and fusional amplitude training after surgery, with only orthoptic treatment in the 5 th case. In all cases, fusion area improvement was recognized during training. In 2 cases examined by cine mode MRI before and after surgery, we observed improved eye movement after training, the effectiveness of which was thereby proven. Also, using cine mode MRI we were able to determine the character of incarcerated tissue and the cause of eye movement disturbance. We conclude that it blowout fracture, cine mode MRI may be useful in selecting treatment and observing its effectiveness. (author)

  20. Cine-Club

    CERN Multimedia

    Cine-Club

    2011-01-01

    Lundi 12 décembre 2011 à 18:30 Salle du Conseil Comme chaque année avant Noël, le CINE-CLUB du CERN est heureux d’inviter petits et grands à une projection gratuite du film : Ponyo sur la falaise (Japon, 2008, Hayao Miyazaki) Le petit Sosuke vit avec sa mère sur une haute falaise surplombant la mer. Un beau jour, il découvre sur la plage caillouteuse Ponyo, une petite fille poisson. Ponyo est si fascinée par Sosuke et le monde terrestre que son désir le plus cher est de devenir un être humain. Mais Fujimoto, son magicien de père, n’est pas du tout d’accord avec cette idée et il la force à regagner les profondeurs de l’océan. Bien décidée à revoir Sosuke, Ponyo s’échappe de sa prison sous-marine, mais ce faisant elle déclenche une immense catastroph...

  1. Cine-Club

    CERN Multimedia

    CineClub

    2014-01-01

      On the occasion of CERN’s 60th anniversary the CERN CineClub will be showing films from all CERN member states Thursday 10 April 2014 at 20:00 CERN Council Chamber   The Bothersome Man     Directed by Jens Lien (Norway, 2006) 95 minutes   Forty-year-old Andreas arrives in a strange city with no memory of how he got there. He is presented with a job, an apartment-even a beautiful girlfriend. But before long, Andreas notices that something is wrong. The people around him seem cut off from any real emotion, and communicate only in superficialities. All this seems to be governed by a shadowy group of technicians, the ominous Caretakers’, who make sure the city runs smoothly. When they find Andreas is not adjusting to his new life, they keep an increasing watch over his activities...”The Bothersome Man” is a fantastic fable, a parable for modern society’s consumerism and obsession with ap...

  2. CERN CINE CLUB

    CERN Multimedia

    CERN CINE CLUB

    2010-01-01

    CERN CINE CLUB Thursday 24 June 2010 at 20:30 / Jeudi 24 Juin 2010 à 20:30 CERN Main Auditorium / Amphithéâtre Principal   The Raggedy Rawney By/de : Bob Hoskins (UK, 1988) - 104 min With/avec: Dexter Fletcher, Bob Hoskins, David Hill, Zoe Nathanson, Zoe Wanamaker Bob Hoskins makes his directorial debut with this lyrical, mystical fable about the strength of family and the transcendence of love. When a young military recruit named Tom goes AWOL after his first taste of battle, he must disguise himself with face paint and women’s clothing to avoid being captured by his vengeful commanding officer. He is instead discovered and taken in by Darky, the leader of an eccentric group of traveling gypsies, who thinks he is a «rawney», a half-mad, half-magical woman who brings good fortune. But when Tom begins a love affair with Darky’s daughter, he sets off a chain reaction of events that will soon put all of their lives in grave ...

  3. Cine Club - Special Event

    CERN Multimedia

    Cine Club

    2017-01-01

    Special event on Thursday 4 May 2017 at 18:30 CERN Council Chamber In collaboration with the CERN Running Club and the Women In Technology initiative, the CERN CineClub is happy to announce the screening of the film Free to Run Directed by Pierre Morath Switzerland, 2016, 99 minutes Today, all anybody needs to run is the determination and a pair of the right shoes. But just fifty years ago, running was viewed almost exclusively as the domain of elite male athletes who competed on tracks. With insight and propulsive energy, director Pierre Morath traces running's rise to the 1960s, examining how the liberation movements and newfound sense of personal freedom that defined the era took the sport out of the stadiums and onto the streets, and how legends like Steve Prefontaine, Fred Lebow, and Kathrine Switzer redefined running as a populist phenomenon. Original version French; English subtitles. http://freetorun.ch/ Come along to watch the film and learn more about the history of popular races and amat...

  4. Cine en la biblioteca

    Directory of Open Access Journals (Sweden)

    Iñaki Esteban Bilbao

    2015-01-01

    Full Text Available Comunicación y Género, acopia trabajos que vinculan a la mujer y la comunicación y sus esfuerzos por no ser consideradas inferiores. Del libro a la pantalla, acerca a algunas reflexiones sobre el proceso que lleva del libro a la pantalla por la polémica que despierta la adaptación de las novelas, puesto que muchos señalan que la trama se ve desfigurada. Se agregan varios temas sobre el cine. En "Días de Radio", se plantea el que a pesar de la censura y los problemas económicos que deben enfrentar, las radios populares de América Latina, siguen ganando espacios, de su desarrollo depende una mayor profundización de la democracia. Tema de debate es Estrategias comunicativas en Chiapas y Haití. Se añaden dos entrevistas a un caricaturista brasileño y a un corresponsal de guerra muy prestigiado.

  5. Movimientos migratorios y cine

    Directory of Open Access Journals (Sweden)

    Manuel Galiano León

    2009-12-01

    Full Text Available Nuestro artículo se ha centrado en el estudio de las películas de mayor distribución comercial para ver cómo éstas han representado el tema de la migración y qué reflexiones sugieren. En general, creemos que el reconocimiento por parte del cine de este fenómeno es positivo y que indudablemente constituye una fuente esencial para comprender las realidades migratorias distinguiendo discursos diversos y explicitando intenciones y hechos que intentan ocultarse u olvidarse.__________________ABSTRACT:Our article focuses on the study of the most commercial movies to check the way they present the subject of immigration and which kind of reflections they suggest. In general, we believe that the recognition of the migration phenomena in the cinema world is positive and it is actually an esential source to understand the reality of migration, being able to recognize different discourses and showing intentions and facts which are hidden or forgotten.  

  6. Comparison of fluoro and cine coronary angiography: balancing acceptable outcomes with a reduction in radiation dose.

    Science.gov (United States)

    Olcay, Ayhan; Guler, Ekrem; Karaca, Ibrahim Oguz; Omaygenc, Mehmet Onur; Kizilirmak, Filiz; Olgun, Erkam; Yenipinar, Esra; Cakmak, Huseyin Altug; Duman, Dursun

    2015-04-01

    Use of last fluoro hold (LFH) mode in fluoroscopy, which enables the last live image to be saved and displayed, could reduce radiation during percutaneous coronary intervention when compared with cine mode. No previous study compared coronary angiography radiation doses and image quality between LFH and conventional cine mode techniques. We compared cumulative dose-area product (DAP), cumulative air kerma, fluoroscopy time, contrast use, interobserver variability of visual assessment between LFH angiography, and conventional cine angiography techniques. Forty-six patients were prospectively enrolled into the LFH group and 82 patients into the cine angiography group according to operator decision. Mean cumulative DAP was higher in the cine group vs the LFH group (50058.98 ± 53542.71 mGy•cm² vs 11349.2 ± 8796.46 mGy•cm²; Pcine group vs the LFH group (3.87 ± 5.08 minutes vs 1.66 ± 1.51 minutes; Pcine group vs the LFH group (112.07 ± 43.79 cc vs 88.15 ± 23.84 cc; Pcine and LFH angiography groups (0.66680 ± 0.19309 vs 0.54193 ± 0.31046; P=.20). Radiation doses, contrast use, and fluoroscopy times are lower in fluoroscopic LFH angiography vs cine angiography. Interclass variability of visual stenosis estimation between three operators was not different between cine and LFH groups. Fluoroscopic LFH images conventionally have inferior diagnostic quality when compared with cine coronary angiography, but with new angiographic systems with improved LFH image quality, these images may be adequate for diagnostic coronary angiography.

  7. MO-FG-202-01: A Fast Yet Sensitive EPID-Based Real-Time Treatment Verification System

    International Nuclear Information System (INIS)

    Ahmad, M; Nourzadeh, H; Neal, B; Siebers, J; Watkins, W

    2016-01-01

    Purpose: To create a real-time EPID-based treatment verification system which robustly detects treatment delivery and patient attenuation variations. Methods: Treatment plan DICOM files sent to the record-and-verify system are captured and utilized to predict EPID images for each planned control point using a modified GPU-based digitally reconstructed radiograph algorithm which accounts for the patient attenuation, source energy fluence, source size effects, and MLC attenuation. The DICOM and predicted images are utilized by our C++ treatment verification software which compares EPID acquired 1024×768 resolution frames acquired at ∼8.5hz from Varian Truebeam™ system. To maximize detection sensitivity, image comparisons determine (1) if radiation exists outside of the desired treatment field; (2) if radiation is lacking inside the treatment field; (3) if translations, rotations, and magnifications of the image are within tolerance. Acquisition was tested with known test fields and prior patient fields. Error detection was tested in real-time and utilizing images acquired during treatment with another system. Results: The computational time of the prediction algorithms, for a patient plan with 350 control points and 60×60×42cm^3 CT volume, is 2–3minutes on CPU and <27 seconds on GPU for 1024×768 images. The verification software requires a maximum of ∼9ms and ∼19ms for 512×384 and 1024×768 resolution images, respectively, to perform image analysis and dosimetric validations. Typical variations in geometric parameters between reference and the measured images are 0.32°for gantry rotation, 1.006 for scaling factor, and 0.67mm for translation. For excess out-of-field/missing in-field fluence, with masks extending 1mm (at isocenter) from the detected aperture edge, the average total in-field area missing EPID fluence was 1.5mm2 the out-of-field excess EPID fluence was 8mm^2, both below error tolerances. Conclusion: A real-time verification software, with

  8. MO-FG-202-01: A Fast Yet Sensitive EPID-Based Real-Time Treatment Verification System

    Energy Technology Data Exchange (ETDEWEB)

    Ahmad, M; Nourzadeh, H; Neal, B; Siebers, J [University of Virginia Health System, Charlottesville, VA (United States); Watkins, W

    2016-06-15

    Purpose: To create a real-time EPID-based treatment verification system which robustly detects treatment delivery and patient attenuation variations. Methods: Treatment plan DICOM files sent to the record-and-verify system are captured and utilized to predict EPID images for each planned control point using a modified GPU-based digitally reconstructed radiograph algorithm which accounts for the patient attenuation, source energy fluence, source size effects, and MLC attenuation. The DICOM and predicted images are utilized by our C++ treatment verification software which compares EPID acquired 1024×768 resolution frames acquired at ∼8.5hz from Varian Truebeam™ system. To maximize detection sensitivity, image comparisons determine (1) if radiation exists outside of the desired treatment field; (2) if radiation is lacking inside the treatment field; (3) if translations, rotations, and magnifications of the image are within tolerance. Acquisition was tested with known test fields and prior patient fields. Error detection was tested in real-time and utilizing images acquired during treatment with another system. Results: The computational time of the prediction algorithms, for a patient plan with 350 control points and 60×60×42cm^3 CT volume, is 2–3minutes on CPU and <27 seconds on GPU for 1024×768 images. The verification software requires a maximum of ∼9ms and ∼19ms for 512×384 and 1024×768 resolution images, respectively, to perform image analysis and dosimetric validations. Typical variations in geometric parameters between reference and the measured images are 0.32°for gantry rotation, 1.006 for scaling factor, and 0.67mm for translation. For excess out-of-field/missing in-field fluence, with masks extending 1mm (at isocenter) from the detected aperture edge, the average total in-field area missing EPID fluence was 1.5mm2 the out-of-field excess EPID fluence was 8mm^2, both below error tolerances. Conclusion: A real-time verification software, with

  9. A simple method to back-project isocenter dose of radiotherapy treatments using EPID transit dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Silveira, T.B.; Cerbaro, B.Q.; Rosa, L.A.R. da, E-mail: thiago.fisimed@gmail.com, E-mail: tbsilveira@inca.gov.br [Instituto de Radioproteção e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro - RJ (Brazil)

    2017-07-01

    The aim of this work was to implement a simple algorithm to evaluate isocenter dose in a phantom using the back-projected transmitted dose acquired using an Electronic Portal Imaging Device (EPID) available in a Varian Trilogy accelerator with two nominal 6 and 10 MV photon beams. This algorithm was developed in MATLAB language, to calibrate EPID measured dose in absolute dose, using a deconvolution process, and to incorporate all scattering and attenuation contributions due to photon interactions with phantom. Modeling process was simplified by using empirical curve adjustments to describe the contribution of scattering and attenuation effects. The implemented algorithm and method were validated employing 19 patient treatment plans with 104 clinical irradiation fields projected on the phantom used. Results for EPID absolute dose calibration by deconvolution have showed percent deviations lower than 1%. Final method validation presented average percent deviations between isocenter doses calculated by back-projection and isocenter doses determined with ionization chamber of 1,86% (SD of 1,00%) and -0,94% (SD of 0,61%) for 6 and 10 MV, respectively. Normalized field by field analysis showed deviations smaller than 2% for 89% of all data for 6 MV beams and 94% for 10 MV beams. It was concluded that the proposed algorithm possesses sufficient accuracy to be used for in vivo dosimetry, being sensitive to detect dose delivery errors bigger than 3-4% for conformal and intensity modulated radiation therapy techniques. (author)

  10. SU-G-TeP4-07: Automatic EPID-Based 2D Measurement of MLC Leaf Offset as a Quality Control Tool

    Energy Technology Data Exchange (ETDEWEB)

    Ritter, T; Moran, J [The University of Michigan, Ann Arbor, MI (United States); Schultz, B [University of Michigan, Ann Arbor, MI (United States); Kim, G [University of California, San Diego, La Jolla, CA (United States); Barnes, M [Calvary Mater Hospital Newcastle, Warratah, NSW (Australia); Perez, M [North Sydney Cancer Center, Sydney (Australia); Farrey, K [University of Chicago, Chicago, IL (United States); Popple, R [University Alabama Birmingham, Birmingham, AL (United States); Greer, P [Calvary Mater Newcastle, Newcastle (Australia)

    2016-06-15

    Purpose: The MLC dosimetric leaf gap (DLG) and transmission are measured parameters which impact the dosimetric accuracy of IMRT and VMAT plans. This investigation aims to develop an efficient and accurate routine constancy check of the physical DLG in two dimensions. Methods: The manufacturer’s recommended DLG measurement method was modified by using 5 fields instead of 11 and by utilizing the Electronic Portal Imaging Device (EPID). Validations were accomplished using an ion chamber (IC) in solid water and a 2D IC array. EPID data was collected for 6 months on multiple TrueBeam linacs using both Millennium and HD MLCs at 5 different clinics in an international consortium. Matlab code was written to automatically analyze the images and calculate the 2D results. Sensitivity was investigated by introducing deliberate leaf position errors. MLC calibration and initialization history was recorded to allow quantification of their impact. Results were analyzed using statistical process control (SPC). Results: The EPID method took approximately 5 minutes. Due to detector response, the EPID measured DLG and transmission differed from the IC values but were reproducible and consistent with changes measured using the ICs. For the Millennium MLC, the EPID measured DLG and transmission were both consistently lower than IC results. The EPID method was implemented as leaf offset and transmission constancy tests (LOC and TC). Based on 6 months of measurements, the initial leaf-specific action thresholds for changes from baseline were set to 0.1 mm. Upper and lower control limits for variation were developed for each machine. Conclusion: Leaf offset and transmission constancy tests were implemented on Varian HD and Millennium MLCs using an EPID and found to be efficient and accurate. The test is effective for monitoring MLC performance using dynamic delivery and performing process control on the DLG in 2D, thus enhancing dosimetric accuracy. This work was supported by a grant

  11. Evaluation of left ventricular myocardial function in patients with coronary artery disease and myocardial dysfunction before and after coronary artery bypass grafting by cine magnetic resonance imaging

    International Nuclear Information System (INIS)

    Juergens, K.U.; Renger, B.; Reimer, P.; Tombach, B.; Heindel, W.; Wichter, T.; Bruns, H.J.; Vahlhaus, C.; Janssen, F.W.; Breithardt, G.

    2001-01-01

    Purpose: To evaluate left ventricular (LV) myocardial function in ten patients with coronary artery disease (CAD) preoperatively and 6 months after coronary bypass grafting (CABG) by cardiac MRI. Material and methods: Ten patients (mean 65.2±5.9 years) with angiographically proven CAD and an indication for elective CABG underwent prospective evaluation of global LV function and regional wall motion by Cine-MRI at rest using a multiphase FLASH-2D sequence following regions of interest (ROI)-defined diagnostics of regional myocardial wall motion by means of levocardiography. Within the ROIs a total of 613 LV myocardial segments were analyzed preceding and following surgical revascularization. Results were compared with the data of 10 healthy volunteers. Results: Preoperatively, patients showed reduced stroke volume and ejection fraction compared with volunteers (p [de

  12. SU-F-T-469: A Clinically Observed Discrepancy Between Image-Based and Log- Based MLC Position

    Energy Technology Data Exchange (ETDEWEB)

    Neal, B; Ahmed, M; Siebers, J [University of Virginia Health System, Charlottesville, VA (United States)

    2016-06-15

    Purpose: To present a clinical case which challenges the base assumption of log-file based QA, by showing that the actual position of a MLC leaf can suddenly deviate from its programmed and logged position by >1 mm as observed with real-time imaging. Methods: An EPID-based exit-fluence dosimetry system designed to prevent gross delivery errors was used in cine mode to capture portal images during treatment. Visual monitoring identified an anomalous MLC leaf pair gap not otherwise detected by the automatic position verification. The position of the erred leaf was measured on EPID images and log files were analyzed for the treatment in question, the prior day’s treatment, and for daily MLC test patterns acquired on those treatment days. Additional standard test patterns were used to quantify the leaf position. Results: Whereas the log file reported no difference between planned and recorded positions, image-based measurements showed the leaf to be 1.3±0.1 mm medial from the planned position. This offset was confirmed with the test pattern irradiations. Conclusion: It has been clinically observed that log-file derived leaf positions can differ from their actual positions by >1 mm, and therefore cannot be considered to be the actual leaf positions. This cautions the use of log-based methods for MLC or patient quality assurance without independent confirmation of log integrity. Frequent verification of MLC positions through independent means is a necessary precondition to trusting log file records. Intra-treatment EPID imaging provides a method to capture departures from MLC planned positions. Work was supported in part by Varian Medical Systems.

  13. Motion Correction using Coil Arrays (MOCCA) for Free-Breathing Cardiac Cine MRI

    Science.gov (United States)

    Hu, Peng; Hong, Susie; Moghari, Mehdi H.; Goddu, Beth; Goepfert, Lois; Kissinger, Kraig V.; Hauser, Thomas H.; Manning, Warren J; Nezafat, Reza

    2014-01-01

    In this study, we present a motion compensation technique based on coil arrays (MOCCA) and evaluate its application in free-breathing respiratory self-gated cine MRI. MOCCA takes advantages of the fact that motion-induced changes in k-space signal are modulated by individual coil sensitivity profiles. In the proposed implementation of MOCCA self-gating for free-breathing cine MRI, the k-space center line is acquired at the beginning of each k-space segment for each cardiac cycle with 4 repetitions. For each k-space segment, the k-space center line acquired immediately before was used to select one of the 4 acquired repetitions to be included in the final self-gated cine image by calculating the cross-correlation between the k-space center line with a reference line. The proposed method was tested on a cohort of healthy adult subjects for subjective image quality and objective blood-myocardium border sharpness. The method was also tested on a cohort of patients to compare the left and right ventricular volumes and ejection fraction measurements with that of standard breath-hold cine MRI. Our data indicate that the proposed MOCCA method provides significantly improved image quality and sharpness compared to free-breathing cine without respiratory self-gating, and provides similar volume measurements compared with breath-hold cine MRI. PMID:21773986

  14. MO-D-213-08: Remote Dosimetric Credentialing for Clinical Trials with the Virtual EPID Standard Phantom Audit (VESPA)

    International Nuclear Information System (INIS)

    Lehmann, J; Miri, N; Vial, P; Hatton, J; Zwan, B; Sloan, K; Craig, A; Beenstock, V; Molloy, T; Greer, P

    2015-01-01

    Purpose: Report on implementation of a Virtual EPID Standard Phantom Audit (VESPA) for IMRT to support credentialing of facilities for clinical trials. Data is acquired by local facility staff and transferred electronically. Analysis is performed centrally. Methods: VESPA is based on published methods and a clinically established IMRT QA procedure, here extended to multi-vendor equipment. Facilities, provided with web-based comprehensive instructions and CT datasets, create IMRT treatment plans. They deliver the treatments directly to their EPID without phantom or couch in the beam. They also deliver a set of simple calibration fields. Collected EPID images are uploaded electronically. In the analysis, the dose is projected back into a virtual phantom and 3D gamma analysis is performed. 2D dose planes and linear dose profiles can be analysed when needed for clarification. Results: Pilot facilities covering a range of planning and delivery systems have performed data acquisition and upload successfully. Analysis showed agreement comparable to local experience with the method. Advantages of VESPA are (1) fast turnaround mainly driven by the facility’s capability to provide the requested EPID images, (2) the possibility for facilities performing the audit in parallel, as there is no need to wait for a phantom, (3) simple and efficient credentialing for international facilities, (4) a large set of data points, and (5) a reduced impact on resources and environment as there is no need to transport heavy phantoms or audit staff. Limitations of the current implementation of VESPA for trials credentialing are that it does not provide absolute dosimetry, therefore a Level 1 audit still required, and that it relies on correctly delivered open calibration fields, which are used for system calibration. Conclusion: The implemented EPID based IMRT audit system promises to dramatically improve credentialing efficiency for clinical trials and wider applications. VESPA for VMAT

  15. MO-D-213-08: Remote Dosimetric Credentialing for Clinical Trials with the Virtual EPID Standard Phantom Audit (VESPA)

    Energy Technology Data Exchange (ETDEWEB)

    Lehmann, J [Calvary Mater Newcastle, Newcastle, NSW (Australia); University of Sydney, Sydney, NSW (Australia); Miri, N [University of Newcastle, Newcastle, NSW (Australia); Vial, P [Liverpool Hospital, Liverpool, NSW (Australia); Hatton, J [Trans Tasman Radiation Oncology Group (TROG), Newcastle, NSW (Australia); Zwan, B; Sloan, K [Gosford Hospital, Gosford, NSW (Australia); Craig, A; Beenstock, V [Canterbury Regional Cancer and Haematology Service, Christchurch (New Zealand); Molloy, T [Orange Hospital, Orange, NSW (Australia); Greer, P [Calvary Mater Newcastle, Newcastle, NSW (Australia); University of Newcastle, Newcastle, NSW (Australia)

    2015-06-15

    Purpose: Report on implementation of a Virtual EPID Standard Phantom Audit (VESPA) for IMRT to support credentialing of facilities for clinical trials. Data is acquired by local facility staff and transferred electronically. Analysis is performed centrally. Methods: VESPA is based on published methods and a clinically established IMRT QA procedure, here extended to multi-vendor equipment. Facilities, provided with web-based comprehensive instructions and CT datasets, create IMRT treatment plans. They deliver the treatments directly to their EPID without phantom or couch in the beam. They also deliver a set of simple calibration fields. Collected EPID images are uploaded electronically. In the analysis, the dose is projected back into a virtual phantom and 3D gamma analysis is performed. 2D dose planes and linear dose profiles can be analysed when needed for clarification. Results: Pilot facilities covering a range of planning and delivery systems have performed data acquisition and upload successfully. Analysis showed agreement comparable to local experience with the method. Advantages of VESPA are (1) fast turnaround mainly driven by the facility’s capability to provide the requested EPID images, (2) the possibility for facilities performing the audit in parallel, as there is no need to wait for a phantom, (3) simple and efficient credentialing for international facilities, (4) a large set of data points, and (5) a reduced impact on resources and environment as there is no need to transport heavy phantoms or audit staff. Limitations of the current implementation of VESPA for trials credentialing are that it does not provide absolute dosimetry, therefore a Level 1 audit still required, and that it relies on correctly delivered open calibration fields, which are used for system calibration. Conclusion: The implemented EPID based IMRT audit system promises to dramatically improve credentialing efficiency for clinical trials and wider applications. VESPA for VMAT

  16. El poder en el cine

    OpenAIRE

    Brisset Mart??n, Demetrio E.

    2008-01-01

    Desde la antropolog??a visual, se aborda la faceta del Poder relativa a sus diversas manifestaciones. Entre ellas, elegimos el cine, especialmente su g??nero de drama judicial, y analizamos comparativamente El proceso f??lmico de Welles y la novela de Kafka.

  17. Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience

    International Nuclear Information System (INIS)

    Zijtveld, Mathilda van; Dirkx, Maarten L.P.; Boer, Hans C.J. de; Heijmen, Ben J.M.

    2006-01-01

    Background and purpose: In our clinic a QA program for IMRT verification, fully based on dosimetric measurements with electronic portal imaging devices (EPID), has been running for over 3 years. The program includes a pre-treatment dosimetric check of all IMRT fields. During a complete treatment simulation at the linac, a portal dose image (PDI) is acquired with the EPID for each patient field and compared with a predicted PDI. In this paper, the results of this pre-treatment procedure are analysed, and intercepted errors are reported. An automated image analysis procedure is proposed to limit the number of fields that need human intervention in PDI comparison. Materials and methods: Most of our analyses are performed using the γ index with 3% local dose difference and 3 mm distance to agreement as reference values. Scalar parameters are derived from the γ values to summarize the agreement between measured and predicted 2D PDIs. Areas with all pixels having γ values larger than one are evaluated, making decisions based on clinically relevant criteria more straightforward. Results: In 270 patients, the pre-treatment checks revealed four clinically relevant errors. Calculation of statistics for a group of 75 patients showed that the patient-averaged mean γ value inside the field was 0.43 ± 0.13 (1 SD) and only 6.1 ± 6.8% of pixels had a γ value larger than one. With the proposed automated image analysis scheme, visual inspection of images can be avoided in 2/3 of the cases. Conclusion: EPIDs may be used for high accuracy and high resolution routine verification of IMRT fields to intercept clinically relevant dosimetric errors prior to the start of treatment. For the majority of fields, PDI comparison can fully rely on an automated procedure, avoiding excessive workload

  18. [Visualization of laryngopharynx during swallowing of negative contrast media (air) with cine mode 64-row MDCT].

    Science.gov (United States)

    Fudeya, Taku; Otake, Shoichiro; Watabe, Hirotaka; Mitsuoka, Takashi; Yoshikawa, Akitoshi

    2010-05-20

    Video fluoroscopic examination of swallowing generally needs a contrast media such as a barium sulfate. Since the examination is usually performed in patients with dysphasia, there is a risk of aspiration. We tried to visualize the laryngopharynx during swallowing of negative contrast media (air) with 64-row multi-detector computed tomography (64-MDCT). Cine mode 64-MDCT was performed to visualize the laryngopharynx in 4 healthy volunteers during swallowing of negative contrast media (air). The data were converted to three-dimensional (3D) images of 2 conditions (air and bone) and sagittal images of the soft tissue condition at a workstation. These images were sent to a personal computer and modeled to 3D cine images with Digital Imaging and Communication in Medicine (DICOM) Viewer and Quick Time Player. 3D cine images demonstrated movements of the epiglottis, vallecula, piriform sinus, tongue, pharyngeal wall, hyoid bone and thyroid cartilage.

  19. Feasibility of vocal fold abduction and adduction assessment using cine-MRI.

    Science.gov (United States)

    Baki, Marina Mat; Menys, Alex; Atkinson, David; Bassett, Paul; Morley, Simon; Beale, Timothy; Sandhu, Guri; Naduvilethil, Georgekutty; Stevenson, Nicola; Birchall, Martin A; Punwani, Shonit

    2017-02-01

    Determine feasibility of vocal fold (VF) abduction and adduction assessment by cine magnetic resonance imaging (cine-MRI) METHODS: Cine-MRI of the VF was performed on five healthy and nine unilateral VF paralysis (UVFP) participants using an axial gradient echo acquisition with temporal resolution of 0.7 s. VFs were continuously imaged with cine-MRI during a 10-s period of quiet respiration and phonation. Scanning was repeated twice within an individual session and then once again at a 1-week interval. Asymmetry of VF position during phonation (VF phonation asymmetry, VFPa) and respiration (VF respiration asymmetry, VFRa) was determined. Percentage reduction in total glottal area between respiration and phonation (VF abduction potential, VFAP) was derived to measure overall mobility. An un-paired t-test was used to compare differences between groups. Intra-session, inter-session and inter-reader repeatability of the quantitative metrics was evaluated using intraclass correlation coefficient (ICC). VF position asymmetry (VFPa and VFRa) was greater (p=0.012; p=0.001) and overall mobility (VFAP) was lower (p=0.008) in UVFP patients compared with healthy participants. ICC of repeatability of all metrics was good, ranged from 0.82 to 0.95 except for the inter-session VFPa (0.44). Cine-MRI is feasible for assessing VF abduction and adduction. Derived quantitative metrics have good repeatability. • Cine-MRI is used to assess vocal folds (VFs) mobility: abduction and adduction. • New quantitative metrics are derived from VF position and abduction potential. • Cine-MRI able to depict the difference between normal and abnormal VF mobility. • Cine-MRI derived quantitative metrics have good repeatability.

  20. Commissioning of Portal Dosimetry and characterization of an EPID

    International Nuclear Information System (INIS)

    Olbi, D.S.; Sales, C.P.; Nakandakari, M.V.N.

    2016-01-01

    The development of technologies compensator blocks, MLC, high dose rate accelerators, treatment planning systems, among others, permitted that new treatment techniques in radiotherapy were created. Such techniques have the capacity to modulate radiation beam fluency (IMRT, VMAT), or to deliver high doses in few fractions or unique fractions (SRS). Following the same tendency, quality control of planning became more complex. It is necessary to evaluate the fluency delivered by the accelerator. Its levels of does and its spatial distribution should co-occur with the fluency calculated by TPS. Acquisition of new detector devices in quality control of treatments is fundamental to apply techniques. Portal Vision is a device EPID has the capacity to operate either in image mode or dosimetry mode, with the allowance of Portal Dosimetry. To evaluated planning in IMRT, the device is irradiated using planning e, therefore, the fluency measured is compared with calculated fluency, through gamma analysis. The aim of this work was to perform tests of commissioning of this device. (author)

  1. Comparison of left and right ventricular ejection and filling parameters by fast cine MR imaging in breath-hold technique: clinical study of 42 patients with cardiomyopathy and coronary heart disease

    International Nuclear Information System (INIS)

    Rominger, M.B.; Bachmann, G.F.; Geuer, M.; Puzik, M.; Rau, W.S.; Ricken, W.W.

    1999-01-01

    Purpose: Quantification of left and right ventricular filling and ejection of localized and diffuse heart diseases with fast cine MR imaging in breath-hold technique. Methods: 42 patients (14 idiopathic dilated cardiomyopathies (DCM), 13 hypertrophic cardiomyopathies (HCM) and 15 coronary artery diseases (CAD)) and 10 healthy volunteers were examined. Time-volume-curves of three left ventricular and one right ventricular slices were evaluated and peak ejection and filling rates (PER, PFR end-diastolic volume (EDV)/s) time to PER and PFR (TPER, TPFR ms) and time of end-systole (TSYS in % RR-intervall) were calculated. Results: There were significant regional and left-/right-sided differences of the filling and ejection of both ventricles within and between the different groups. In DCM the left ventricular PFR was reduced (DCM 3.1 EDV/s; volunteers 4.9 EDV/s) and Z-SYS prolonged (DCM 50.1%; volunteers 35.4%). In CAD there were localized decreased filling rates in comparison to the normal volunteer group (left ventricle: basal: 2.9 and 6.3 EDV/s, apical: 4.4 and 6.3 EDV/s; right ventricle: 3.6 and 5.7 EDV/s). HCM typically showed an isovolumetric lengthening of the endsystole. Conclusions: Cardiac MR imaging in breath-hold technique is suitable for measuring contraction and relaxation disturbances of localized and diffuse heart diseases by means of ejection and filling volume indices. (orig.) [de

  2. Automated x-ray/light field congruence using the LINAC EPID panel

    Energy Technology Data Exchange (ETDEWEB)

    Polak, Wojciech [Department of Medical Physics, Royal Surrey County Hospital, Guildford GU2 7XX (United Kingdom); Department of Medical Physics, Radiotherapy Section, Queen Alexandra Hospital NHS Trust, Portsmouth PO6 3LY (United Kingdom); O' Doherty, Jim [Division of Imaging Sciences and Biomedical Engineering, King' s College London, London SE1 7EH, United Kingdom and Department of Medical Physics, Royal Surrey County Hospital, Guildford GU2 7XX (United Kingdom); Jones, Matt [Department of Medical Physics, Royal Surrey County Hospital, Guildford GU2 7XX (United Kingdom)

    2013-03-15

    Purpose: X-ray/light field alignment is a test described in many guidelines for the routine quality control of clinical linear accelerators (LINAC). Currently, the gold standard method for measuring alignment is through utilization of radiographic film. However, many modern LINACs are equipped with an electronic portal imaging device (EPID) that may be used to perform this test and thus subsequently reducing overall cost, processing, and analysis time, removing operator dependency and the requirement to sustain the departmental film processor. Methods: This work describes a novel method of utilizing the EPID together with a custom inhouse designed jig and automatic image processing software allowing measurement of the light field size, x-ray field size, and congruence between them. The authors present results of testing the method for aS1000 and aS500 Varian EPID detectors for six LINACs at a range of energies (6, 10, and 15 MV) in comparison with the results obtained from the use of radiographic film. Results: Reproducibility of the software in fully automatic operation under a range of operating conditions for a single image showed a congruence of 0.01 cm with a coefficient of variation of 0. Slight variation in congruence repeatability was noted through semiautomatic processing by four independent operators due to manual marking of positions on the jig. Testing of the methodology using the automatic method shows a high precision of 0.02 mm compared to a maximum of 0.06 mm determined by film processing. Intraindividual examination of operator measurements of congruence was shown to vary as much as 0.75 mm. Similar congruence measurements of 0.02 mm were also determined for a lower resolution EPID (aS500 model), after rescaling of the image to the aS1000 image size. Conclusions: The designed methodology was proven to be time efficient, cost effective, and at least as accurate as using the gold standard radiographic film. Additionally, congruence testing can be

  3. Automated x-ray/light field congruence using the LINAC EPID panel

    International Nuclear Information System (INIS)

    Polak, Wojciech; O’Doherty, Jim; Jones, Matt

    2013-01-01

    Purpose: X-ray/light field alignment is a test described in many guidelines for the routine quality control of clinical linear accelerators (LINAC). Currently, the gold standard method for measuring alignment is through utilization of radiographic film. However, many modern LINACs are equipped with an electronic portal imaging device (EPID) that may be used to perform this test and thus subsequently reducing overall cost, processing, and analysis time, removing operator dependency and the requirement to sustain the departmental film processor. Methods: This work describes a novel method of utilizing the EPID together with a custom inhouse designed jig and automatic image processing software allowing measurement of the light field size, x-ray field size, and congruence between them. The authors present results of testing the method for aS1000 and aS500 Varian EPID detectors for six LINACs at a range of energies (6, 10, and 15 MV) in comparison with the results obtained from the use of radiographic film. Results: Reproducibility of the software in fully automatic operation under a range of operating conditions for a single image showed a congruence of 0.01 cm with a coefficient of variation of 0. Slight variation in congruence repeatability was noted through semiautomatic processing by four independent operators due to manual marking of positions on the jig. Testing of the methodology using the automatic method shows a high precision of 0.02 mm compared to a maximum of 0.06 mm determined by film processing. Intraindividual examination of operator measurements of congruence was shown to vary as much as 0.75 mm. Similar congruence measurements of 0.02 mm were also determined for a lower resolution EPID (aS500 model), after rescaling of the image to the aS1000 image size. Conclusions: The designed methodology was proven to be time efficient, cost effective, and at least as accurate as using the gold standard radiographic film. Additionally, congruence testing can be

  4. Automated x-ray/light field congruence using the LINAC EPID panel.

    Science.gov (United States)

    Polak, Wojciech; O'Doherty, Jim; Jones, Matt

    2013-03-01

    X-ray/light field alignment is a test described in many guidelines for the routine quality control of clinical linear accelerators (LINAC). Currently, the gold standard method for measuring alignment is through utilization of radiographic film. However, many modern LINACs are equipped with an electronic portal imaging device (EPID) that may be used to perform this test and thus subsequently reducing overall cost, processing, and analysis time, removing operator dependency and the requirement to sustain the departmental film processor. This work describes a novel method of utilizing the EPID together with a custom inhouse designed jig and automatic image processing software allowing measurement of the light field size, x-ray field size, and congruence between them. The authors present results of testing the method for aS1000 and aS500 Varian EPID detectors for six LINACs at a range of energies (6, 10, and 15 MV) in comparison with the results obtained from the use of radiographic film. Reproducibility of the software in fully automatic operation under a range of operating conditions for a single image showed a congruence of 0.01 cm with a coefficient of variation of 0. Slight variation in congruence repeatability was noted through semiautomatic processing by four independent operators due to manual marking of positions on the jig. Testing of the methodology using the automatic method shows a high precision of 0.02 mm compared to a maximum of 0.06 mm determined by film processing. Intraindividual examination of operator measurements of congruence was shown to vary as much as 0.75 mm. Similar congruence measurements of 0.02 mm were also determined for a lower resolution EPID (aS500 model), after rescaling of the image to the aS1000 image size. The designed methodology was proven to be time efficient, cost effective, and at least as accurate as using the gold standard radiographic film. Additionally, congruence testing can be easily performed for all four cardinal

  5. Commissioning of Portal Dosimetry and characterization of an EPID; Comissionamento de Portal Dosimetry e caracterizacao de EPID

    Energy Technology Data Exchange (ETDEWEB)

    Olbi, D.S.; Sales, C.P. [Universidade de Sao Paulo (USP), SP (Brazil). Faculdade de Medicina; Nakandakari, M.V.N., E-mail: diego.olbi@hc.fm.usp.br [Instituto do Cancer do Estado de Sao Paulo, SP (Brazil). Servico de Radioterapia

    2016-07-01

    The development of technologies compensator blocks, MLC, high dose rate accelerators, treatment planning systems, among others, permitted that new treatment techniques in radiotherapy were created. Such techniques have the capacity to modulate radiation beam fluency (IMRT, VMAT), or to deliver high doses in few fractions or unique fractions (SRS). Following the same tendency, quality control of planning became more complex. It is necessary to evaluate the fluency delivered by the accelerator. Its levels of does and its spatial distribution should co-occur with the fluency calculated by TPS. Acquisition of new detector devices in quality control of treatments is fundamental to apply techniques. Portal Vision is a device EPID has the capacity to operate either in image mode or dosimetry mode, with the allowance of Portal Dosimetry. To evaluated planning in IMRT, the device is irradiated using planning e, therefore, the fluency measured is compared with calculated fluency, through gamma analysis. The aim of this work was to perform tests of commissioning of this device. (author)

  6. Feasibility of free-breathing, GRAPPA-based, real-time cardiac cine assessment of left-ventricular function in cardiovascular patients at 3 T

    International Nuclear Information System (INIS)

    Zhu, Xiaomei; Schwab, Felix; Marcus, Roy; Hetterich, Holger; Theisen, Daniel; Kramer, Harald; Notohamiprodjo, Mike; Schlett, Christopher L.; Nikolaou, Konstantin; Reiser, Maximilian F.

    2015-01-01

    Highlights: • Grappa-based real time cine cardiac MRI is feasible for assessment of left ventricular function. • Significant underestimation of systolic function, peak ejection and filling rates needs to be considered. • Heart rate is the only positive predictor of the deviation of obtained parameters. - Abstract: Objectives: To determine the feasibility of free-breathing, GRAPPA-based, real-time (RT) cine 3 T cardiac magnetic resonance imaging (MRI) with high acceleration factors for the assessment of left-ventricular function in a cohort of patients as compared to conventional segmented cine imaging. Materials and methods: In this prospective cohort study, subjects with various cardiac conditions underwent MRI involving two RT cine sequences (high resolution and low resolution) and standard segmented cine imaging. Standard qualitative and quantitative parameters of left-ventricular function were quantified. Results: Among 25 subjects, 24 were included in the analysis (mean age: 50.5 ± 21 years, 67% male, 25% with cardiomyopathy). RT cine derived quantitative parameters of volumes and left ventricular mass were strongly correlated with segmented cine imaging (intraclass correlation coefficient [ICC]: >0.72 for both RT cines) but correlation for peak ejection and filling rates were moderate to poor for both RT cines (ICC < 0.40). Similarly, RT cines significantly underestimated peak ejection and filling rates (>103.2 ± 178 ml/s). Among patient-related factors, heart rate was strongly predictive for deviation of measurements (p < 0.05). Conclusions: RT cine MRI at 3 T is feasible for qualitative and quantitative assessment of left ventricular function for low and high-resolution sequences but results in significant underestimation of systolic function, peak ejection and filling rates

  7. Verification of PTV margins for IMRT prostate cancer using EPID; Verificacao das margens de PTV para IMRT de cancer de prostata utilizando EPID

    Energy Technology Data Exchange (ETDEWEB)

    Leidens, Matheus; Santos, Romulo R.; Estacio, Daniela R. [Pontificia Universidade Catolica do Rio Grande do Sul (PUCRS), Porto Alegre, RS (Brazil). Hospital Sao Lucas. Servico de Fisica Medica; Silva, Ana Maria Marques da, E-mail: matheus_leidens@hotmail.com [Pontificia Universidade Catolica do Rio Grande do Sul (PUCRS), Porto Alegre, RS (Brazil). Faculdade de Fisica

    2014-12-15

    The aim of this work is to present the results of a strategy to define the PTV margins for patients with prostate cancer treated with IMRT technique, due to geometrical uncertainties associated with the planned placement. 341 images of 31 patients in supine position, before applying the fractions, were obtained using an EPID attached to a linear accelerator, where only setup errors were studied. The displacements were analyzed in relation to the AP (antero-posterior), SI (superior-inferior) and LR (left-right) directions. The distribution pattern of systematic displacement deviation values were 0.12 cm, 0.06 cm, 0.02 cm and the standard deviation of the distribution of random deviations was 0.62 cm, 0.53 cm, and 0.24 cm in the AP, SI and LR directions, respectively. Data evaluation, according to Stroom and Heijmen’s method, suggests that PTV margins should be 0.66 cm in the AP direction, 0.49 cm in the SI direction and 0.20 cm in the LR direction. These data show a high reproducibility in the positioning of patients, given by a method for the correction of planned relative to the bony anatomy checked with the EPID position. (author)

  8. Holistic segmentation of the lung in cine MRI.

    Science.gov (United States)

    Kovacs, William; Hsieh, Nathan; Roth, Holger; Nnamdi-Emeratom, Chioma; Bandettini, W Patricia; Arai, Andrew; Mankodi, Ami; Summers, Ronald M; Yao, Jianhua

    2017-10-01

    Duchenne muscular dystrophy (DMD) is a childhood-onset neuromuscular disease that results in the degeneration of muscle, starting in the extremities, before progressing to more vital areas, such as the lungs. Respiratory failure and pneumonia due to respiratory muscle weakness lead to hospitalization and early mortality. However, tracking the disease in this region can be difficult, as current methods are based on breathing tests and are incapable of distinguishing between muscle involvements. Cine MRI scans give insight into respiratory muscle movements, but the images suffer due to low spatial resolution and poor signal-to-noise ratio. Thus, a robust lung segmentation method is required for accurate analysis of the lung and respiratory muscle movement. We deployed a deep learning approach that utilizes sequence-specific prior information to assist the segmentation of lung in cine MRI. More specifically, we adopt a holistically nested network to conduct image-to-image holistic training and prediction. One frame of the cine MRI is used in the training and applied to the remainder of the sequence ([Formula: see text] frames). We applied this method to cine MRIs of the lung in the axial, sagittal, and coronal planes. Characteristic lung motion patterns during the breathing cycle were then derived from the segmentations and used for diagnosis. Our data set consisted of 31 young boys, age [Formula: see text] years, 15 of whom suffered from DMD. The remaining 16 subjects were age-matched healthy volunteers. For validation, slices from inspiratory and expiratory cycles were manually segmented and compared with results obtained from our method. The Dice similarity coefficient for the deep learning-based method was [Formula: see text] for the sagittal view, [Formula: see text] for the axial view, and [Formula: see text] for the coronal view. The holistic neural network approach was compared with an approach using Demon's registration and showed superior performance. These

  9. Feasibility of free-breathing, GRAPPA-based, real-time cardiac cine assessment of left-ventricular function in cardiovascular patients at 3T.

    Science.gov (United States)

    Zhu, Xiaomei; Schwab, Felix; Marcus, Roy; Hetterich, Holger; Theisen, Daniel; Kramer, Harald; Notohamiprodjo, Mike; Schlett, Christopher L; Nikolaou, Konstantin; Reiser, Maximilian F; Bamberg, Fabian

    2015-05-01

    To determine the feasibility of free-breathing, GRAPPA-based, real-time (RT) cine 3T cardiac magnetic resonance imaging (MRI) with high acceleration factors for the assessment of left-ventricular function in a cohort of patients as compared to conventional segmented cine imaging. In this prospective cohort study, subjects with various cardiac conditions underwent MRI involving two RT cine sequences (high resolution and low resolution) and standard segmented cine imaging. Standard qualitative and quantitative parameters of left-ventricular function were quantified. Among 25 subjects, 24 were included in the analysis (mean age: 50.5±21 years, 67% male, 25% with cardiomyopathy). RT cine derived quantitative parameters of volumes and left ventricular mass were strongly correlated with segmented cine imaging (intraclass correlation coefficient [ICC]: >0.72 for both RT cines) but correlation for peak ejection and filling rates were moderate to poor for both RT cines (ICCcines significantly underestimated peak ejection and filling rates (>103.2±178 ml/s). Among patient-related factors, heart rate was strongly predictive for deviation of measurements (pcine MRI at 3T is feasible for qualitative and quantitative assessment of left ventricular function for low and high-resolution sequences but results in significant underestimation of systolic function, peak ejection and filling rates. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  10. Dose calculation with respiration-averaged CT processed from cine CT without a respiratory surrogate

    International Nuclear Information System (INIS)

    Riegel, Adam C.; Ahmad, Moiz; Sun Xiaojun; Pan Tinsu

    2008-01-01

    Dose calculation for thoracic radiotherapy is commonly performed on a free-breathing helical CT despite artifacts caused by respiratory motion. Four-dimensional computed tomography (4D-CT) is one method to incorporate motion information into the treatment planning process. Some centers now use the respiration-averaged CT (RACT), the pixel-by-pixel average of the ten phases of 4D-CT, for dose calculation. This method, while sparing the tedious task of 4D dose calculation, still requires 4D-CT technology. The authors have recently developed a means to reconstruct RACT directly from unsorted cine CT data from which 4D-CT is formed, bypassing the need for a respiratory surrogate. Using RACT from cine CT for dose calculation may be a means to incorporate motion information into dose calculation without performing 4D-CT. The purpose of this study was to determine if RACT from cine CT can be substituted for RACT from 4D-CT for the purposes of dose calculation, and if increasing the cine duration can decrease differences between the dose distributions. Cine CT data and corresponding 4D-CT simulations for 23 patients with at least two breathing cycles per cine duration were retrieved. RACT was generated four ways: First from ten phases of 4D-CT, second, from 1 breathing cycle of images, third, from 1.5 breathing cycles of images, and fourth, from 2 breathing cycles of images. The clinical treatment plan was transferred to each RACT and dose was recalculated. Dose planes were exported at orthogonal planes through the isocenter (coronal, sagittal, and transverse orientations). The resulting dose distributions were compared using the gamma (γ) index within the planning target volume (PTV). Failure criteria were set to 2%/1 mm. A follow-up study with 50 additional lung cancer patients was performed to increase sample size. The same dose recalculation and analysis was performed. In the primary patient group, 22 of 23 patients had 100% of points within the PTV pass γ criteria

  11. Right ventricular volume estimation with cine MRI

    International Nuclear Information System (INIS)

    Sawachika, Takashi

    1993-01-01

    To quantitate right ventricular (RV) volumes easily using cine MRI, we developed a new method called 'modified area-length method (MOAL method)'. To validate this method, we compared it to the conventional Simpson's rule. Magnetom H15 (Siemens) was used and 6 normal volunteers and 21 patients with various RV sizes were imaged with ECG triggered gradient echo method (FISP, TR 50 ms, TE 12 ms, slice thickness 9 mm). For Simpson's rule transverse images of 12 sequential views which cover whole heart were acquired. For the MOAL method, two orthogonal views were imaged. One was the sagittal view which includes RV outflow tract and the other was the coronal view defined from the sagittal image to cover the whole RV. From these images the area (As, Ac) of RV and the longest distance between RV apex and pulmonary valve (Lmax) were determined. By correlating RV volumes measured by Simpson's rule to As*Ac/Lmax the RV volume could be estimated as follows: V=0.85*As*Ac/Lmax+4.55. Thus the MOAL method demonstrated excellent accuracy to quantitate RV volume and the acquisition time abbreviated to one fifth compared with Simpson's rule. This should be a highly promising method for routine clinical application. (author)

  12. PA/Lateral chest X-ray is equivalent to cine-fluoroscopy for the detection of conductor externalization in defibrillation leads.

    Science.gov (United States)

    Steinberg, Christian; Sarrazin, Jean-François; Philippon, François; Champagne, Jean; Molin, Franck; Nault, Isabelle; Blier, Louis; Bouchard, Marc-André; Arsenault, Jean; O'Hara, Gilles

    2015-01-01

    Riata™ and Riata ST defibrillation leads (St. Jude Medical, Sylmar, CA, USA) are susceptible to insulation defects with conductor externalization. Cine-fluoroscopy is considered to be the gold standard for the documentation of insulation defects, but similar detection rates have been reported for posterior-anterior (PA)/lateral chest x-ray (CXR) with zooming. Prospective single-center study to assess the diagnostic equivalence of a PA/lateral CXR with zooming for the detection of Riata insulation defects in a direct comparison to cine-fluoroscopy. Seventy-eight consecutive patients underwent 3-view cine-fluoroscopy and a PA/lateral CXR. All CXRs and cine-fluoroscopy images were reviewed by blinded electrophysiologists and staff radiologists. Forty-four of 78 patients had an abnormal cine-fluoroscopy (56%). The diagnostic correlation between PA/lateral CXR and cine-fluoroscopy was excellent (κ = 0.90; 95% confidence interval 0.80-1.00). PA/lateral CXR was equivalent to cine-fluoroscopy for the detection of conductor externalization showing a sensitivity of 97.7% and a specificity of 91.2%. The mean radiation effective dose of CXR was significantly lower compared to cine-fluoroscopy (0.09 millisievert [mSV] vs 0.85 ± 0.47 mSv; P cine-fluoroscopy for the detection of Riata insulation defects and should be considered as the preferred screening method. ©2014 Wiley Periodicals, Inc.

  13. EPID-based in vivo dosimetry for stereotactic body radiotherapy of non-small cell lung tumors: Initial clinical experience.

    Science.gov (United States)

    Consorti, R; Fidanzio, A; Brainovich, V; Mangiacotti, F; De Spirito, M; Mirri, M A; Petrucci, A

    2017-10-01

    EPID-based in vivo dosimetry (IVD) has been implemented for stereotactic body radiotherapy treatments of non-small cell lung cancer to check both isocenter dose and the treatment reproducibility comparing EPID portal images. 15 patients with lung tumors of small dimensions and treated with volumetric modulated arc therapy were enrolled for this initial experience. IVD tests supplied ratios R between in vivo reconstructed and planned isocenter doses. Moreover a γ-like analysis between daily EPID portal images and a reference one, in terms of percentage of points with γ-value smaller than 1, P γlevels of 5% for R ratio, P γlevel, and an average P γ90%. Paradigmatic discrepancies were observed in three patients: a set-up error and a patient morphological change were identified thanks to CBCT image analysis whereas the third discrepancy was not fully justified. This procedure can provide improved patient safety as well as a first step to integrate IVD and CBCT dose recalculation. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  14. Miércoles al cine

    OpenAIRE

    Aguado Franco, Juan Carlos

    2014-01-01

    Se analiza un caso concreto de demanda ante una iniciativa empresarial: los miércoles al cine. Se analiza un caso concreto de demanda ante una iniciativa empresarial: los miércoles al cine. Fundamentos del Análisis Económico

  15. Comparison of standard, prone and cine MRI in the evaluation of tethered cord

    International Nuclear Information System (INIS)

    Singh, Sukhjinder; Kline-Fath, Beth; Racadio, Judy M.; Bierbrauer, Karin; Salisbury, Shelia; Macaluso, Maurizio; Jackson, Elizabeth C.; Egelhoff, John C.

    2012-01-01

    Tethered cord syndrome (TCS) is defined by abnormal traction on the spinal cord that confines its movement. Surgical cord release usually stops neurological deterioration; therefore, early and accurate neuroradiological diagnosis is important. Supine MRI is the imaging modality of choice, but prone MRI and cine MRI can demonstrate cord movement. We compared the diagnostic accuracies of standard MRI, prone MRI and cine MRI in patients with clinical suspicion of TCS and evaluated inter-reader reliability for MR imaging. Children who underwent MRI for suspicion of TCS were retrospectively identified. Supine, prone and cine MRI studies were re-read by two pediatric neuroradiologists. Conus level, filum appearance and cord movement were documented. Thirteen of 49 children had tethered cord documented at surgery. Conus level had the highest diagnostic accuracy (sensitivity 69-77%, specificity 94%, positive predictive value 82-83%, negative predictive value 89-92%, correct diagnosis 88-90%) and highest between-reader concordance (98%). Prone and cine MRI did not add to the accuracy of the supine imaging. Conus level provides the highest diagnostic accuracy and inter-reader reliability in TCS. Until a larger series is evaluated, it remains questionable whether prone or cine MRI provides enough additional diagnostic information to warrant routine use. (orig.)

  16. Comparison of standard, prone and cine MRI in the evaluation of tethered cord

    Energy Technology Data Exchange (ETDEWEB)

    Singh, Sukhjinder [Cohen Children' s Medical Center, Department of Radiology, New Hyde Park, NY (United States); Kline-Fath, Beth; Racadio, Judy M. [Cincinnati Children' s Hospital Medical Center, Department of Radiology, Cincinnati, OH (United States); Bierbrauer, Karin [Cincinnati Children' s Hospital Medical Center, Department of Neurosurgery, Cincinnati, OH (United States); Salisbury, Shelia; Macaluso, Maurizio [Cincinnati Children' s Hospital Medical Center, Division of Biostatistics and Epidemiology, Cincinnati, OH (United States); Jackson, Elizabeth C. [Cincinnati Children' s Hospital Medical Center, Department of Pediatrics, Division of Nephrology, Cincinnati, OH (United States); Egelhoff, John C. [Cincinnati Children' s Hospital Medical Center, Department of Radiology, Cincinnati, OH (United States); Phoenix Children' s Hospital, Department of Radiology, Phoenix, AZ (United States)

    2012-06-15

    Tethered cord syndrome (TCS) is defined by abnormal traction on the spinal cord that confines its movement. Surgical cord release usually stops neurological deterioration; therefore, early and accurate neuroradiological diagnosis is important. Supine MRI is the imaging modality of choice, but prone MRI and cine MRI can demonstrate cord movement. We compared the diagnostic accuracies of standard MRI, prone MRI and cine MRI in patients with clinical suspicion of TCS and evaluated inter-reader reliability for MR imaging. Children who underwent MRI for suspicion of TCS were retrospectively identified. Supine, prone and cine MRI studies were re-read by two pediatric neuroradiologists. Conus level, filum appearance and cord movement were documented. Thirteen of 49 children had tethered cord documented at surgery. Conus level had the highest diagnostic accuracy (sensitivity 69-77%, specificity 94%, positive predictive value 82-83%, negative predictive value 89-92%, correct diagnosis 88-90%) and highest between-reader concordance (98%). Prone and cine MRI did not add to the accuracy of the supine imaging. Conus level provides the highest diagnostic accuracy and inter-reader reliability in TCS. Until a larger series is evaluated, it remains questionable whether prone or cine MRI provides enough additional diagnostic information to warrant routine use. (orig.)

  17. Megavoltage conebeam CT cine as final verification of treatment plan in lung stereotactic body radiotherapy.

    Science.gov (United States)

    Kudithipudi, Vijay; Gayou, Olivier; Colonias, Athanasios

    2016-06-01

    To analyse the clinical impact of megavoltage conebeam computed tomography (MV-CBCT) cine on internal target volume (ITV) coverage in lung stereotactic body radiotherapy (SBRT). One hundred and six patients received lung SBRT. All underwent 4D computed tomography simulation followed by treatment via image guided 3D conformal or intensity modulated radiation. Prior to SBRT, all patients underwent MV-CBCT cine, in which raw projections are displayed as beam's-eye-view fluoroscopic series with the planning target volume (PTV) projected onto each image, enabling verification of tumour motion relative to the PTV and assessment of adequacy of treatment margin. Megavoltage conebeam computed tomography cine was completed 1-2 days prior to SBRT. Four patients (3.8%) had insufficient ITV coverage inferiorly at cine review. All four plans were changed by adding 5 mm on the PTV margin inferiorly. The mean change in PTV volumes was 3.9 cubic centimetres (cc) (range 1.85-6.32 cc). Repeat cine was performed after plan modification to ensure adequate PTV coverage in the modified plans. PTV margin was adequate in the majority of patients with this technique. MV-CBCT cine did show insufficient coverage in a small subset of patients. Insufficient PTV margins may be a function of 4D CT simulation inadequacies or deficiencies in visualizing the ITV inferior border in the full-inhale phase. MV-CBCT cine is a valuable tool for final verification of PTV margins. © 2016 The Royal Australian and New Zealand College of Radiologists.

  18. Amorphous silicon EPID calibration for dosimetric applications: comparison of a method based on Monte Carlo prediction of response with existing techniques

    International Nuclear Information System (INIS)

    Parent, L; Fielding, A L; Dance, D R; Seco, J; Evans, P M

    2007-01-01

    For EPID dosimetry, the calibration should ensure that all pixels have a similar response to a given irradiation. A calibration method (MC), using an analytical fit of a Monte Carlo simulated flood field EPID image to correct for the flood field image pixel intensity shape, was proposed. It was compared with the standard flood field calibration (FF), with the use of a water slab placed in the beam to flatten the flood field (WS) and with a multiple field calibration where the EPID was irradiated with a fixed 10 x 10 field for 16 different positions (MF). The EPID was used in its normal configuration (clinical setup) and with an additional 3 mm copper slab (modified setup). Beam asymmetry measured with a diode array was taken into account in MC and WS methods. For both setups, the MC method provided pixel sensitivity values within 3% of those obtained with the MF and WS methods (mean difference 2 ) and IMRT fields to within 3% of that obtained with WS and MF calibrations while differences with images calibrated with the FF method for fields larger than 10 x 10 cm 2 were up to 8%. MC, WS and MF methods all provided a major improvement on the FF method. Advantages and drawbacks of each method were reviewed

  19. SU-G-TeP2-01: Can EPID Based Measurement Replace Traditional Daily Output QA On Megavoltage Linac?

    International Nuclear Information System (INIS)

    Saleh, Z; Tang, X; Song, Y; Obcemea, C; Beeban, N; Chan, M; Li, X; Tang, G; Lim, S; Lovelock, D; LoSasso, T; Mechalakos, J; Both, S

    2016-01-01

    Purpose: To investigate the long term stability and viability of using EPID-based daily output QA via in-house and vendor driven protocol, to replace conventional QA tools and improve QA efficiency. Methods: Two Varian TrueBeam machines (TB1&TB2) equipped with electronic portal imaging devices (EPID) were employed in this study. Both machines were calibrated per TG-51 and used clinically since Oct 2014. Daily output measurement for 6/15 MV beams were obtained using SunNuclear DailyQA3 device as part of morning QA. In addition, in-house protocol was implemented for EPID output measurement (10×10 cm fields, 100 MU, 100cm SID, output defined over an ROI of 2×2 cm around central axis). Moreover, the Varian Machine Performance Check (MPC) was used on both machines to measure machine output. The EPID and DailyQA3 based measurements of the relative machine output were compared and cross-correlated with monthly machine output as measured by an A12 Exradin 0.65cc Ion Chamber (IC) serving as ground truth. The results were correlated using Pearson test. Results: The correlations among DailyQA3, in-house EPID and Varian MPC output measurements, with the IC for 6/15 MV were similar for TB1 (0.83–0.95) and TB2 (0.55–0.67). The machine output for the 6/15MV beams on both machines showed a similar trend, namely an increase over time as indicated by all measurements, requiring a machine recalibration after 6 months. This drift is due to a known issue with pressurized monitor chamber which tends to leak over time. MPC failed occasionally but passed when repeated. Conclusion: The results indicate that the use of EPID for daily output measurements has the potential to become a viable and efficient tool for daily routine LINAC QA, thus eliminating weather (T,P) and human setup variability and increasing efficiency of the QA process.

  20. SU-G-TeP2-01: Can EPID Based Measurement Replace Traditional Daily Output QA On Megavoltage Linac?

    Energy Technology Data Exchange (ETDEWEB)

    Saleh, Z; Tang, X; Song, Y; Obcemea, C; Beeban, N; Chan, M; Li, X; Tang, G; Lim, S; Lovelock, D; LoSasso, T; Mechalakos, J; Both, S [Memorial Sloan-Kettering Cancer Center, NY (United States)

    2016-06-15

    Purpose: To investigate the long term stability and viability of using EPID-based daily output QA via in-house and vendor driven protocol, to replace conventional QA tools and improve QA efficiency. Methods: Two Varian TrueBeam machines (TB1&TB2) equipped with electronic portal imaging devices (EPID) were employed in this study. Both machines were calibrated per TG-51 and used clinically since Oct 2014. Daily output measurement for 6/15 MV beams were obtained using SunNuclear DailyQA3 device as part of morning QA. In addition, in-house protocol was implemented for EPID output measurement (10×10 cm fields, 100 MU, 100cm SID, output defined over an ROI of 2×2 cm around central axis). Moreover, the Varian Machine Performance Check (MPC) was used on both machines to measure machine output. The EPID and DailyQA3 based measurements of the relative machine output were compared and cross-correlated with monthly machine output as measured by an A12 Exradin 0.65cc Ion Chamber (IC) serving as ground truth. The results were correlated using Pearson test. Results: The correlations among DailyQA3, in-house EPID and Varian MPC output measurements, with the IC for 6/15 MV were similar for TB1 (0.83–0.95) and TB2 (0.55–0.67). The machine output for the 6/15MV beams on both machines showed a similar trend, namely an increase over time as indicated by all measurements, requiring a machine recalibration after 6 months. This drift is due to a known issue with pressurized monitor chamber which tends to leak over time. MPC failed occasionally but passed when repeated. Conclusion: The results indicate that the use of EPID for daily output measurements has the potential to become a viable and efficient tool for daily routine LINAC QA, thus eliminating weather (T,P) and human setup variability and increasing efficiency of the QA process.

  1. Cine y turismo: intersecciones polivalentes.

    OpenAIRE

    Rey-Reguillo, Antonia del

    2012-01-01

    Con el objeto de servir de introducción a los tres artículos que componen el dossier sobre Cine y Turismo contenido en la revista, este texto reflexiona someramente sobre el amplio abanico de intersecciones existente entre las industrias cinematográfica y turística ajustándose al caso concreto de la producción fílmica española considerada en tres etapas concretas: los documentales etnoturísticos surgidos en el período histórico correspondiente a la Segunda República, el documental 'España ins...

  2. Homosexualidad latente en el cine

    OpenAIRE

    Sánchez del Pulgar Legido, Rosa Mª

    2015-01-01

    Desde el principio del siglo XX en los Estados Unidos y Europa, las personas vivían su homosexualidad a escondidas por temor a las leyes que la castigaban; el cine pues, les representa del mismo modo creando una subcultura en la que pueden ser ellos mismos. La cinematografía clásica y los años posteriores se componen de numerosos filmes cargados de representaciones homosexuales de manera oculta. Interpretados desde una lectura queer, conoceremos las mil maneras de sugerir a los gais y a las l...

  3. Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle

    International Nuclear Information System (INIS)

    Clarke, M F; Budgell, G J

    2008-01-01

    Amorphous silicon electronic portal imaging devices (EPIDs) are used to perform routine quality control (QC) checks on the multileaf collimators (MLCs) at this centre. Presently, these checks are performed at gantry angle 0 0 and are considered to be valid for all other angles. Since therapeutic procedures regularly require the delivery of MLC-defined fields to the patient at a wide range of gantry angles, the accuracy of the QC checks at other gantry angles has been investigated. When the gantry is rotated to angles other than 0 0 it was found that the apparent pixel size measured using the EPID varies up to a maximum value of 0.0015 mm per pixel due to a sag in the EPID of up to 9.2 mm. A correction factor was determined using two independent methods at a range of gantry angles between 0 deg. and 360 deg. The EPID was used to measure field sizes (defined by both x-jaws and MLC) at a range of gantry angles and, after this correction had been applied, any residual gravitational sag was studied. It was found that, when fields are defined by the x-jaws and y-back-up jaws, no errors of greater than 0.5 mm were measured and that these errors were no worse when the MLC was used. It was therefore concluded that, provided the correction is applied, measurements of the field size are, in practical terms, unaffected by gantry angle. Experiments were also performed to study how the reproducibility of individual leaves is affected by gantry angle. Measurements of the relative position of each individual leaf (minor offsets) were performed at a range of gantry angles and repeated three times. The position reproducibility was defined by the RMS error in the position of each leaf and this was found to be 0.24 mm and 0.21 mm for the two leaf banks at a gantry angle of 0 0 . When measurements were performed at a range of gantry angles, these reproducibility values remained within 0.09 mm and 0.11 mm. It was therefore concluded that the calibration of the Elekta MLC is stable at

  4. Compressed sensing reconstruction of cardiac cine MRI using golden angle spiral trajectories.

    Science.gov (United States)

    Tolouee, Azar; Alirezaie, Javad; Babyn, Paul

    2015-11-01

    In dynamic cardiac cine Magnetic Resonance Imaging (MRI), the spatiotemporal resolution is limited by the low imaging speed. Compressed sensing (CS) theory has been applied to improve the imaging speed and thus the spatiotemporal resolution. The purpose of this paper is to improve CS reconstruction of under sampled data by exploiting spatiotemporal sparsity and efficient spiral trajectories. We extend k-t sparse algorithm to spiral trajectories to achieve high spatio temporal resolutions in cardiac cine imaging. We have exploited spatiotemporal sparsity of cardiac cine MRI by applying a 2D+time wavelet-Fourier transform. For efficient coverage of k-space, we have used a modified version of multi shot (interleaved) spirals trajectories. In order to reduce incoherent aliasing artifact, we use different random undersampling pattern for each temporal frame. Finally, we have used nonuniform fast Fourier transform (NUFFT) algorithm to reconstruct the image from the non-uniformly acquired samples. The proposed approach was tested in simulated and cardiac cine MRI data. Results show that higher acceleration factors with improved image quality can be obtained with the proposed approach in comparison to the existing state-of-the-art method. The flexibility of the introduced method should allow it to be used not only for the challenging case of cardiac imaging, but also for other patient motion where the patient moves or breathes during acquisition. Copyright © 2015 Elsevier Inc. All rights reserved.

  5. Feasibility of vocal fold abduction and adduction assessment using cine-MRI

    International Nuclear Information System (INIS)

    Baki, Marina Mat; Menys, Alex; Morley, Simon; Beale, Timothy; Atkinson, David; Punwani, Shonit; Bassett, Paul; Sandhu, Guri; Naduvilethil, Georgekutty; Stevenson, Nicola; Birchall, Martin A.

    2017-01-01

    Determine feasibility of vocal fold (VF) abduction and adduction assessment by cine magnetic resonance imaging (cine-MRI) Cine-MRI of the VF was performed on five healthy and nine unilateral VF paralysis (UVFP) participants using an axial gradient echo acquisition with temporal resolution of 0.7 s. VFs were continuously imaged with cine-MRI during a 10-s period of quiet respiration and phonation. Scanning was repeated twice within an individual session and then once again at a 1-week interval. Asymmetry of VF position during phonation (VF phonation asymmetry, VFPa) and respiration (VF respiration asymmetry, VFRa) was determined. Percentage reduction in total glottal area between respiration and phonation (VF abduction potential, VFAP) was derived to measure overall mobility. An un-paired t-test was used to compare differences between groups. Intra-session, inter-session and inter-reader repeatability of the quantitative metrics was evaluated using intraclass correlation coefficient (ICC). VF position asymmetry (VFPa and VFRa) was greater (p=0.012; p=0.001) and overall mobility (VFAP) was lower (p=0.008) in UVFP patients compared with healthy participants. ICC of repeatability of all metrics was good, ranged from 0.82 to 0.95 except for the inter-session VFPa (0.44). Cine-MRI is feasible for assessing VF abduction and adduction. Derived quantitative metrics have good repeatability. (orig.)

  6. Feasibility of vocal fold abduction and adduction assessment using cine-MRI

    Energy Technology Data Exchange (ETDEWEB)

    Baki, Marina Mat [National University of Malaysia, Faculty of Medicine, Kuala Lumpur (Malaysia); Menys, Alex; Morley, Simon; Beale, Timothy [University College London, Centre for Medical Imaging, London (United Kingdom); Atkinson, David; Punwani, Shonit [University College London, Centre for Medical Imaging, London (United Kingdom); Royal National Throat Nose Ear Hospital, University College London Hospital NHS Trust, London (United Kingdom); Bassett, Paul [University College London, London (United Kingdom); Sandhu, Guri [Charing Cross Hospital, Imperial College Healthcare NHS Trust, London (United Kingdom); Naduvilethil, Georgekutty; Stevenson, Nicola [Royal National Throat Nose Ear Hospital, University College London Hospital NHS Trust, London (United Kingdom); Birchall, Martin A. [Royal National Throat Nose Ear Hospital, University College London Hospital NHS Trust, London (United Kingdom); University of California, Davis, Department of Otolaryngology, Davis, CA (United States); University College London, Ear Institute, London (United Kingdom)

    2017-02-15

    Determine feasibility of vocal fold (VF) abduction and adduction assessment by cine magnetic resonance imaging (cine-MRI) Cine-MRI of the VF was performed on five healthy and nine unilateral VF paralysis (UVFP) participants using an axial gradient echo acquisition with temporal resolution of 0.7 s. VFs were continuously imaged with cine-MRI during a 10-s period of quiet respiration and phonation. Scanning was repeated twice within an individual session and then once again at a 1-week interval. Asymmetry of VF position during phonation (VF phonation asymmetry, VFPa) and respiration (VF respiration asymmetry, VFRa) was determined. Percentage reduction in total glottal area between respiration and phonation (VF abduction potential, VFAP) was derived to measure overall mobility. An un-paired t-test was used to compare differences between groups. Intra-session, inter-session and inter-reader repeatability of the quantitative metrics was evaluated using intraclass correlation coefficient (ICC). VF position asymmetry (VFPa and VFRa) was greater (p=0.012; p=0.001) and overall mobility (VFAP) was lower (p=0.008) in UVFP patients compared with healthy participants. ICC of repeatability of all metrics was good, ranged from 0.82 to 0.95 except for the inter-session VFPa (0.44). Cine-MRI is feasible for assessing VF abduction and adduction. Derived quantitative metrics have good repeatability. (orig.)

  7. Development of a software of VMAT delivery using EPID

    International Nuclear Information System (INIS)

    Arumugam, Sankar; Xing, Aitang; Jameson, Michael; Holloway, Lois; Goozee, Gary

    2011-01-01

    Full text: Volumetric Modulated Arc Therapy (VMAT) is more complex than standard IMRT, requiring new methodology to evaluate delivery accuracy. Here, we present the development of methodology and a software tool to perform control point based verification of VMAT delivery using an EPID. Individual segment dose comparison allows the verification of VMAT deli very accuracy for individual control-points. An in-house software tool was developed to predict the individual segment dose as measured by EPID for Pinnacle (Philips) generated VMAT plans. The VMAT plans were delivered using an Elekta-synergy accelerator and the segment doses were measured using EPID. A normalised dose comparison of measured and predicted doses was performed using gamma analysis with 3% dose tolerance and 3 mm DTA. The sensitivity of the proposed methodology in detecting delivery errors was studied by delivering a standard intensity pattern with a preset dose error of 4 and 5% in two of its eight control-points. Four clinical plans were also tested using this methodology. The developed software accurately predicts the EPID dose by considering all possible leaf trajectories in VMAT delivery. The mean gamma value and percentage of pixels exceeding the gamma tolerance for a segment with and without delivery errors are shown in Table. From the tabulated values it is evident that the proposed methodology is sensitive in detecting delivery errors above 3% tolerance level. The clinical plans were successfully validated showing a maximum 2.5% of pixels exceeding gamma tolerance. Methodology and a software were successfully developed to perform control-point validation of VMAT delivery using an EPID. Set error in Delivery (%) Mean gamma value% of pixels exceeding Gamma tolerance 0 0.40 1.240.5417.050.6221.8.

  8. Comparison of forward- and back-projection in vivo EPID dosimetry for VMAT treatment of the prostate

    Science.gov (United States)

    Bedford, James L.; Hanson, Ian M.; Hansen, Vibeke N.

    2018-01-01

    In the forward-projection method of portal dosimetry for volumetric modulated arc therapy (VMAT), the integrated signal at the electronic portal imaging device (EPID) is predicted at the time of treatment planning, against which the measured integrated image is compared. In the back-projection method, the measured signal at each gantry angle is back-projected through the patient CT scan to give a measure of total dose to the patient. This study aims to investigate the practical agreement between the two types of EPID dosimetry for prostate radiotherapy. The AutoBeam treatment planning system produced VMAT plans together with corresponding predicted portal images, and a total of 46 sets of gantry-resolved portal images were acquired in 13 patients using an iViewGT portal imager. For the forward-projection method, each acquisition of gantry-resolved images was combined into a single integrated image and compared with the predicted image. For the back-projection method, iViewDose was used to calculate the dose distribution in the patient for comparison with the planned dose. A gamma index for 3% and 3 mm was used for both methods. The results were investigated by delivering the same plans to a phantom and repeating some of the deliveries with deliberately introduced errors. The strongest agreement between forward- and back-projection methods is seen in the isocentric intensity/dose difference, with moderate agreement in the mean gamma. The strongest correlation is observed within a given patient, with less correlation between patients, the latter representing the accuracy of prediction of the two methods. The error study shows that each of the two methods has its own distinct sensitivity to errors, but that overall the response is similar. The forward- and back-projection EPID dosimetry methods show moderate agreement in this series of prostate VMAT patients, indicating that both methods can contribute to the verification of dose delivered to the patient.

  9. Cine y literatura: Narrativa de la identidad

    OpenAIRE

    Laverde Román, Alejandra; Parra, Martha Ligia; Montoya Giraldo, Alejandra; Uribe Alzate, Yennifer; Tobar Álvarez, Margarita

    2010-01-01

    Este artículo es resultado de la investigación La narrativa literaria y audiovisual en la construcción de la identidad cultural colombiana. Una mirada desde el cine y la literatura: adaptaciones de obras literarias llevadas al cine entre 1920 y 2008-1. En el presente texto se desarrolla un análisis de la relación de cine y literatura en la construcción de la identidad colombiana, a partir del rastreo bibliográfico de la crítica y las historias literarias y cinematográficas a un corpus selecci...

  10. Cine y literatura: narrativa de la identidad

    OpenAIRE

    Román, Alejandra Laverde; Parra, Martha Ligia

    2010-01-01

    Este artículo es resultado de la investigación La narrativa literaria y audiovisual enla construcción de la identidad cultural colombiana. Una mirada desde el cine y laliteratura: adaptaciones de obras literarias llevadas al cine entre 1920 y 2008-1.En el presente texto se desarrolla un análisis de la relación de cine y literatura enla construcción de la identidad colombiana, a partir del rastreo bibliográfico de lacrítica y las historias literarias y cinematográficas a un corpus seleccionado...

  11. Cine Computed Tomography Without Respiratory Surrogate in Planning Stereotactic Radiotherapy for Non-Small-Cell Lung Cancer

    International Nuclear Information System (INIS)

    Riegel, Adam C. B.A.; Chang, Joe Y.; Vedam, Sastry S.; Johnson, Valen; Chi, Pai-Chun Melinda; Pan, Tinsu

    2009-01-01

    Purpose: To determine whether cine computed tomography (CT) can serve as an alternative to four-dimensional (4D)-CT by providing tumor motion information and producing equivalent target volumes when used to contour in radiotherapy planning without a respiratory surrogate. Methods and Materials: Cine CT images from a commercial CT scanner were used to form maximum intensity projection and respiratory-averaged CT image sets. These image sets then were used together to define the targets for radiotherapy. Phantoms oscillating under irregular motion were used to assess the differences between contouring using cine CT and 4D-CT. We also retrospectively reviewed the image sets for 26 patients (27 lesions) at our institution who had undergone stereotactic radiotherapy for Stage I non-small-cell lung cancer. The patients were included if the tumor motion was >1 cm. The lesions were first contoured using maximum intensity projection and respiratory-averaged CT image sets processed from cine CT and then with 4D-CT maximum intensity projection and 10-phase image sets. The mean ratios of the volume magnitude were compared with intraobserver variation, the mean centroid shifts were calculated, and the volume overlap was assessed with the normalized Dice similarity coefficient index. Results: The phantom studies demonstrated that cine CT captured a greater extent of irregular tumor motion than did 4D-CT, producing a larger tumor volume. The patient studies demonstrated that the gross tumor defined using cine CT imaging was similar to, or slightly larger than, that defined using 4D-CT. Conclusion: The results of our study have shown that cine CT is a promising alternative to 4D-CT for stereotactic radiotherapy planning

  12. A quantification of the effectiveness of EPID dosimetry and software-based plan verification systems in detecting incidents in radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Bojechko, Casey; Phillps, Mark; Kalet, Alan; Ford, Eric C., E-mail: eford@uw.edu [Department of Radiation Oncology, University of Washington, 1959 N. E. Pacific Street, Seattle, Washington 98195 (United States)

    2015-09-15

    Purpose: Complex treatments in radiation therapy require robust verification in order to prevent errors that can adversely affect the patient. For this purpose, the authors estimate the effectiveness of detecting errors with a “defense in depth” system composed of electronic portal imaging device (EPID) based dosimetry and a software-based system composed of rules-based and Bayesian network verifications. Methods: The authors analyzed incidents with a high potential severity score, scored as a 3 or 4 on a 4 point scale, recorded in an in-house voluntary incident reporting system, collected from February 2012 to August 2014. The incidents were categorized into different failure modes. The detectability, defined as the number of incidents that are detectable divided total number of incidents, was calculated for each failure mode. Results: In total, 343 incidents were used in this study. Of the incidents 67% were related to photon external beam therapy (EBRT). The majority of the EBRT incidents were related to patient positioning and only a small number of these could be detected by EPID dosimetry when performed prior to treatment (6%). A large fraction could be detected by in vivo dosimetry performed during the first fraction (74%). Rules-based and Bayesian network verifications were found to be complimentary to EPID dosimetry, able to detect errors related to patient prescriptions and documentation, and errors unrelated to photon EBRT. Combining all of the verification steps together, 91% of all EBRT incidents could be detected. Conclusions: This study shows that the defense in depth system is potentially able to detect a large majority of incidents. The most effective EPID-based dosimetry verification is in vivo measurements during the first fraction and is complemented by rules-based and Bayesian network plan checking.

  13. An in vivo dose verification method for SBRT–VMAT delivery using the EPID

    Energy Technology Data Exchange (ETDEWEB)

    McCowan, P. M., E-mail: peter.mccowan@cancercare.mb.ca [Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2 (Canada); Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada); Van Uytven, E.; Van Beek, T.; Asuni, G. [Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada); McCurdy, B. M. C. [Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba R3T 2N2 (Canada); Medical Physics Department, CancerCare Manitoba, 675 McDermot Avenue, Winnipeg, Manitoba R3E 0V9 (Canada); Department of Radiology, University of Manitoba, 820 Sherbrook Street, Winnipeg, Manitoba R3A 1R9 (Canada)

    2015-12-15

    Purpose: Radiation treatments have become increasingly more complex with the development of volumetric modulated arc therapy (VMAT) and the use of stereotactic body radiation therapy (SBRT). SBRT involves the delivery of substantially larger doses over fewer fractions than conventional therapy. SBRT–VMAT treatments will strongly benefit from in vivo patient dose verification, as any errors in delivery can be more detrimental to the radiobiology of the patient as compared to conventional therapy. Electronic portal imaging devices (EPIDs) are available on most commercial linear accelerators (Linacs) and their documented use for dosimetry makes them valuable tools for patient dose verification. In this work, the authors customize and validate a physics-based model which utilizes on-treatment EPID images to reconstruct the 3D dose delivered to the patient during SBRT–VMAT delivery. Methods: The SBRT Linac head, including jaws, multileaf collimators, and flattening filter, were modeled using Monte Carlo methods and verified with measured data. The simulation provides energy spectrum data that are used by their “forward” model to then accurately predict fluence generated by a SBRT beam at a plane above the patient. This fluence is then transported through the patient and then the dose to the phosphor layer in the EPID is calculated. Their “inverse” model back-projects the EPID measured focal fluence to a plane upstream of the patient and recombines it with the extra-focal fluence predicted by the forward model. This estimate of total delivered fluence is then forward projected onto the patient’s density matrix and a collapsed cone convolution algorithm calculates the dose delivered to the patient. The model was tested by reconstructing the dose for two prostate, three lung, and two spine SBRT–VMAT treatment fractions delivered to an anthropomorphic phantom. It was further validated against actual patient data for a lung and spine SBRT–VMAT plan. The

  14. Left Ventricular Function Evaluation on a 3T MR Scanner with Parallel RF Transmission Technique: Prospective Comparison of Cine Sequences Acquired before and after Gadolinium Injection.

    Science.gov (United States)

    Caspar, Thibault; Schultz, Anthony; Schaeffer, Mickaël; Labani, Aïssam; Jeung, Mi-Young; Jurgens, Paul Thomas; El Ghannudi, Soraya; Roy, Catherine; Ohana, Mickaël

    To compare cine MR b-TFE sequences acquired before and after gadolinium injection, on a 3T scanner with a parallel RF transmission technique in order to potentially improve scanning time efficiency when evaluating LV function. 25 consecutive patients scheduled for a cardiac MRI were prospectively included and had their b-TFE cine sequences acquired before and right after gadobutrol injection. Images were assessed qualitatively (overall image quality, LV edge sharpness, artifacts and LV wall motion) and quantitatively with measurement of LVEF, LV mass, and telediastolic volume and contrast-to-noise ratio (CNR) between the myocardium and the cardiac chamber. Statistical analysis was conducted using a Bayesian paradigm. No difference was found before or after injection for the LVEF, LV mass and telediastolic volume evaluations. Overall image quality and CNR were significantly lower after injection (estimated coefficient cine after > cine before gadolinium: -1.75 CI = [-3.78;-0.0305], prob(coef>0) = 0% and -0.23 CI = [-0.49;0.04], prob(coef>0) = 4%) respectively), but this decrease did not affect the visual assessment of LV wall motion (cine after > cine before gadolinium: -1.46 CI = [-4.72;1.13], prob(coef>0) = 15%). In 3T cardiac MRI acquired with parallel RF transmission technique, qualitative and quantitative assessment of LV function can reliably be performed with cine sequences acquired after gadolinium injection, despite a significant decrease in the CNR and the overall image quality.

  15. A Technique for Generating Volumetric Cine MRI (VC-MRI)

    Science.gov (United States)

    Harris, Wendy; Ren, Lei; Cai, Jing; Zhang, You; Chang, Zheng; Yin, Fang-Fang

    2016-01-01

    Purpose To develop a technique to generate on-board volumetric-cine MRI (VC-MRI) using patient prior images, motion modeling and on-board 2D-cine MRI. Methods One phase of a 4D-MRI acquired during patient simulation is used as patient prior images. 3 major respiratory deformation patterns of the patient are extracted from 4D-MRI based on principal-component-analysis. The on-board VC-MRI at any instant is considered as a deformation of the prior MRI. The deformation field is represented as a linear combination of the 3 major deformation patterns. The coefficients of the deformation patterns are solved by the data fidelity constraint using the acquired on-board single 2D-cine MRI. The method was evaluated using both XCAT simulation of lung cancer patients and MRI data from four real liver cancer patients. The accuracy of the estimated VC-MRI was quantitatively evaluated using Volume-Percent-Difference(VPD), Center-of-Mass-Shift(COMS), and target tracking errors. Effects of acquisition orientation, region-of-interest(ROI) selection, patient breathing pattern change and noise on the estimation accuracy were also evaluated. Results Image subtraction of ground-truth with estimated on-board VC-MRI shows fewer differences than image subtraction of ground-truth with prior image. Agreement between profiles in the estimated and ground-truth VC-MRI was achieved with less than 6% error for both XCAT and patient data. Among all XCAT scenarios, the VPD between ground-truth and estimated lesion volumes was on average 8.43±1.52% and the COMS was on average 0.93±0.58mm across all time-steps for estimation based on the ROI region in the sagittal cine images. Matching to ROI in the sagittal view achieved better accuracy when there was substantial breathing pattern change. The technique was robust against noise levels up to SNR=20. For patient data, average tracking errors were less than 2 mm in all directions for all patients. Conclusions Preliminary studies demonstrated the

  16. Cine MRI of the thorax in patients with pectus excavatum

    International Nuclear Information System (INIS)

    Herrmann, K.A.; Zech, C.J.; Schoenberg, S.O.; Reiser, M.F.; Strauss, T.; Hatz, R.

    2006-01-01

    Morphologic and dynamic assessment of respiratory chest kinetics was performed in patients with pectus excavatum deformity (PE) using dynamic MRI: cine MRI. Seven consecutive patients with PE (aged 20.3 years±4.0) and ten healthy volunteers of comparable age underwent real-time cine MRI of the chest during breathing on a 1.5 T MR scanner (Magnetom Sonata, Siemens Medical Systems, Erlangen, Germany) using a standard phased array body coil and a half-Fourier single-shot turbo spin echo sequence (HASTE) for dynamic imaging. During deep inspiration and expiration, single-shot sequences were performed in one slice level over 20 s at a frequency of 1 image/s covering the entire thoracic cage in three orientations. Morphology and chest kinetics in patients with PE were analyzed and compared with normal values, and typical patterns of chest kinetics were noted. Three different types of chest morphology in PE were identified: (1) the generally flattened thoracic cage, (2) the ''tilted'' sternum, and (3) the focally deepened sternum. Three patterns of motion correspond to these morphological types: (1) elevation of the sternum and the anterior thoracic wall, (2) angulated elevation of the parasternal rib cage with persistent deepening of the sternum resembling a ''wing beat'' movement, and (3) increased diaphragmatic movements with limited chest wall dynamics. Cine MRI is an adequate radiation-free diagnostic modality for the dynamic imaging of both chest morphology and chest wall kinetics in patients with PE. The pectus severity index can easily be determined and three typical movement patterns of chest wall kinetics identified. (orig.) [de

  17. Investigations on uncertainties in patient positioning for prostate treatment with EPID

    International Nuclear Information System (INIS)

    Bakai, A.; Nuesslin, F.; Paulsen, F.; Plasswilm, L.; Bamberg, M.

    2002-01-01

    Background: Conformal radiotherapy techniques as used in prostate treatment allow to spare normal tissue by conforming the radiation fields to the shape of the planning target volume (PTV). To be able to fully utilize the advantages of these techniques correct patient positioning is an important prerequisite. This study employing an electronic portal imaging device (EPID) investigated the positioning uncertainties that occur in the pelvic region for different patient positioning devices. Patients and Methods: 15 patients with prostate cancer were irradiated with or without rectal balloon/pelvic mask at a linear accelerator with multileaf collimator (MLC). For each patient multiple portal images were taken from different directions and compared to the digitally reconstructed radiographs (DRRs) of the treatment planning system and to simulation films (Table 1, Figure 1). Results: In spite of different positioning devices, all patients showed comparable total positioning uncertainties of 4.0 mm (lateral), 4.5 mm (cranio-caudal) and 1.7 mm (dorso-ventral). The lateral positioning error was reduced for the pelvic mask patients while the cranio-caudal error increased (Table 2, Figure 2). A systematic and a random component sum up to the total positioning error, and a good estimate of the magnitudes of the two is possible from six to eight portal images (Figure 3). Conclusions: With a small number of portal images it is possible to find out the systematic and random positioning error of a patient. Knowledge of the random error can be used to resize the treatment margin which is clinically relevant since this error differs greatly for different patients (Figure 4). Image analysis with EPID is convenient, yet has some problems. For example, one only gets indirect information on the movement of the ventral rectum wall. The successful operation of positioning devices, although, needs further improvement - especially if one focuses on IMRT. (orig.) [de

  18. TU-C-BRE-10: A Streamlined Approach to EPID Transit Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Morris, B; Fontenot, J [Louisiana State University, Baton Rouge, LA (United States); Mary Bird Perkins Cancer Center, Baton Rouge, LA (United States)

    2014-06-15

    Purpose: To investigate the feasibility of a simple and efficient transit dosimetry method using the electronic portal imaging device (EPID) for dose delivery error detection and prevention. Methods: In the proposed method, 2D reference transit images are generated for comparison with online images acquired during treatment. Reference transit images are generated by convolving through-air EPID measurements of each field with pixel-specific kernels selected from a library of pre-calculated Monte Carlo pencil kernels of varying radiological thickness. The kernel used for each pixel is selected based on the calculated radiological thickness of the patient along a line joining the pixel and the virtual source. The accuracy of the technique was evaluated in flat homogeneous and heterogeneous plastic water phantoms, a heterogeneous cylindrical phantom, and an anthropomorphic head phantom. Gamma criteria of 3%/3 mm was used to quantify the accuracy of the technique for the various cases. Results: An average of 99.9% and 99.7% of the points in the comparison between the measured and predicted images passed a 3%/3mm gamma for the homogeneous and heterogeneous plastic water phantoms, respectively. 97.1% of the points passed for the analysis of the heterogeneous cylindrical phantom. For the anthropomorphic head phantom, an average of 97.8% of points passed the 3%/3mm gamma criteria for all field sizes. Failures were observed primarily in areas of drastic thickness or material changes and at the edges of the fields. Conclusion: The data suggest that the proposed transit dosimetry method is a feasible approach to in vivo dose monitoring. Future research efforts could include implementation for more complex fields and sensitivity testing of the method to setup errors and changes in anatomy. Oncology Data Systems provided partial funding support but did not participate in the collection or analysis of data.

  19. A study on characteristics of X-ray detector for CCD-based EPID

    International Nuclear Information System (INIS)

    Chung, Yong Hyun

    1999-02-01

    The combination of the metal plate/phosphor screen as a x-ray detector with a CCD camera is the most popular detector system among various electronic portal imaging devices (EPIDs). There is a need to optimize the thickness of the metal plate/phosphor screen with high detection efficiency and high spatial resolution for effective transferring of anatomical information. In this study, the thickness dependency on the detection efficiency and the spatial resolution of the metal plate/phosphor screen was investigated by calculation and measurement. The result can be used to determine the optimal thickness of the metal plate as well as of the phosphor screen for the x-ray detector design of therapeutic x-ray imaging and for any specific application. Bremsstrahlung spectrum was calculated by Monte Carlo simulation and by Schiff formula. The detection efficiency was calculated from the total absorbed energy in the phosphor screen using the Monte Carlo simulation and the light output was measured. The spatial resolution, which was defined from the spatial distribution of the absorbed energy, was also calculated and the edge spread function was measured. It was found that the detection efficiency and the spatial resolution were mainly determined by the thickness of metal plate and phosphor screen, respectively. It was also revealed that the detection efficiency and the spatial resolution have trade-off in term of the thickness of the phosphor screen. As the phosphor thickness increases, the detection efficiency increases but the spatial resolution decreases. The curve illustrating the trade-off between the detection efficiency and the spatial resolution of the metal plate/phosphor screen detector is obtained as a function of the phosphor thickness. Based on the calculations, prototype CCD-based EPID was developed and then tested by acquiring phantom images for 6 MV x-ray beam. While, among the captured images, each frame suffered from quantum noise, the frame averaging

  20. TrueFisp versus HASTE sequences in 3T cine MRI: Evaluation of image quality during phonation in patients with velopharyngeal insufficiency

    International Nuclear Information System (INIS)

    Kulinna-Cosentini, Christiane; Czerny, Christian; Weber, Michael; Baumann, Arnulf; Sinko, Klaus

    2016-01-01

    To evaluate the image quality of two fast dynamic magnetic resonance imaging (MRI) sequences: True fast imaging with steady state precession (TrueFisp) was compared with half-Fourier acquired single turbo-spin-echo (HASTE) sequence for the characterization of velopharyngeal insufficiency (VPI) in repaired cleft palate patients. Twenty-two patients (10 female and 12 male; mean age, 17.7 ± 10.6 years; range, 9-31) with suspected VPI underwent 3-T MRI using TrueFisp and HASTE sequences. Imaging was performed in the sagittal plane at rest and during phonation of ''ee'' and ''k'' to assess the velum, tongue, posterior pharyngeal wall and a potential VP closure. The results were analysed independently by one radiologist and one orthodontist. HASTE performed better than TrueFisp for all evaluated items, except the tongue evaluation by the orthodontist during phonation of ''k'' and ''ee''. A statistically significant difference in favour of HASTE was observed in assessing the velum at rest and during phonation of ''k'' and ''ee'', and also in assessing VP closure in both raters (p < 0.05). TrueFisp imaging was twice as fast as HASTE (0.36 vs. 0.75 s/image). Dynamic HASTE images were of superior quality to those obtained with TrueFisp, although TrueFisp imaging was twice as fast. (orig.)

  1. Accelerated cardiac cine MRI using locally low rank and finite difference constraints.

    Science.gov (United States)

    Miao, Xin; Lingala, Sajan Goud; Guo, Yi; Jao, Terrence; Usman, Muhammad; Prieto, Claudia; Nayak, Krishna S

    2016-07-01

    To evaluate the potential value of combining multiple constraints for highly accelerated cardiac cine MRI. A locally low rank (LLR) constraint and a temporal finite difference (FD) constraint were combined to reconstruct cardiac cine data from highly undersampled measurements. Retrospectively undersampled 2D Cartesian reconstructions were quantitatively evaluated against fully-sampled data using normalized root mean square error, structural similarity index (SSIM) and high frequency error norm (HFEN). This method was also applied to 2D golden-angle radial real-time imaging to facilitate single breath-hold whole-heart cine (12 short-axis slices, 9-13s single breath hold). Reconstruction was compared against state-of-the-art constrained reconstruction methods: LLR, FD, and k-t SLR. At 10 to 60 spokes/frame, LLR+FD better preserved fine structures and depicted myocardial motion with reduced spatio-temporal blurring in comparison to existing methods. LLR yielded higher SSIM ranking than FD; FD had higher HFEN ranking than LLR. LLR+FD combined the complimentary advantages of the two, and ranked the highest in all metrics for all retrospective undersampled cases. Single breath-hold multi-slice cardiac cine with prospective undersampling was enabled with in-plane spatio-temporal resolutions of 2×2mm(2) and 40ms. Highly accelerated cardiac cine is enabled by the combination of 2D undersampling and the synergistic use of LLR and FD constraints. Copyright © 2016 Elsevier Inc. All rights reserved.

  2. Accelerated high-frame-rate mouse heart cine-MRI using compressed sensing reconstruction.

    Science.gov (United States)

    Motaal, Abdallah G; Coolen, Bram F; Abdurrachim, Desiree; Castro, Rui M; Prompers, Jeanine J; Florack, Luc M J; Nicolay, Klaas; Strijkers, Gustav J

    2013-04-01

    We introduce a new protocol to obtain very high-frame-rate cinematographic (Cine) MRI movies of the beating mouse heart within a reasonable measurement time. The method is based on a self-gated accelerated fast low-angle shot (FLASH) acquisition and compressed sensing reconstruction. Key to our approach is that we exploit the stochastic nature of the retrospective triggering acquisition scheme to produce an undersampled and random k-t space filling that allows for compressed sensing reconstruction and acceleration. As a standard, a self-gated FLASH sequence with a total acquisition time of 10 min was used to produce single-slice Cine movies of seven mouse hearts with 90 frames per cardiac cycle. Two times (2×) and three times (3×) k-t space undersampled Cine movies were produced from 2.5- and 1.5-min data acquisitions, respectively. The accelerated 90-frame Cine movies of mouse hearts were successfully reconstructed with a compressed sensing algorithm. The movies had high image quality and the undersampling artifacts were effectively removed. Left ventricular functional parameters, i.e. end-systolic and end-diastolic lumen surface areas and early-to-late filling rate ratio as a parameter to evaluate diastolic function, derived from the standard and accelerated Cine movies, were nearly identical. Copyright © 2012 John Wiley & Sons, Ltd.

  3. Assessment of small bowel motility in patients with chronic intestinal pseudo-obstruction using cine-MRI.

    Science.gov (United States)

    Ohkubo, Hidenori; Kessoku, Takaomi; Fuyuki, Akiko; Iida, Hiroshi; Inamori, Masahiko; Fujii, Tetsuro; Kawamura, Harunobu; Hata, Yasuo; Manabe, Noriaki; Chiba, Toshimi; Kwee, Thomas C; Haruma, Ken; Matsuhashi, Nobuyuki; Nakajima, Atsushi; Takahara, Taro

    2013-07-01

    Chronic intestinal pseudo-obstruction (CIPO) is a rare, serious motility disorder, with life-threatening complications over time. However, lack of an established, non-invasive diagnostic method has caused delays in the diagnosis of this intractable disease. Cine-magnetic resonance imaging (MRI) is an emerging technique, with a potential to evaluate the motility of the entire bowel. We compared small bowel motility in healthy volunteers, patients with irritable bowel syndrome (IBS), and those with CIPO, using cine-MRI, and evaluated the usefulness of cine-MRI as a novel diagnostic method for CIPO. Twelve healthy volunteers, IBS patients, and CIPO patients prospectively underwent cine-MRI at 1.5 T. Luminal diameter, contraction ratio, and contraction cycle were measured and compared between the groups. Cine-MRI provided sufficient dynamic images to assess the motility of the entire small bowel. Luminal diameter (mean±s.d.) in CIPO patients was significantly higher than that in healthy volunteers and IBS patients (43.4±14.1, 11.1±1.5, and 10.9±1.9 mm, respectively), and contraction ratio was significantly lower in CIPO patients than that in healthy volunteers and IBS patients (17.1±11.0%, 73.0±9.3%, and 74.6±9.4%, respectively). No significant differences were observed in the contraction cycle. This study is the first to assess the clinical utility of cine-MRI in CIPO patients. Cine-MRI clearly detected contractility impairments in CIPO patients. Cine-MRI is noninvasive, radiation-free, and can directly evaluate the entire small bowel peristalsis, and can detect the affected loops at a glance; therefore, it might be extremely useful for the diagnosis and follow-up of CIPO patients in clinical practice.

  4. Analysis of flow dynamics of main pulmonary artery by cine phase contrast MR angiography

    International Nuclear Information System (INIS)

    Honda, Norinari; Machida, Kikuo; Mamiya, Toshio

    1996-01-01

    Nineteen studies of cine phase contrast MR angiography (PCMRA) of main pulmonary artery (MPA) entered the study. Cine PCMRA was obtained by a 1.5T MR imager with a gradient echo sequence coupled with velocity encoding bipolar pulses. Retrospective EKG gating was used. Mean velocity, maximum velocity, and maximum flow rate of MPA were 6.2 to 28 cm/s (mean/SD 13/5.1, n=18), 61 to 148 cm/s (mean/SD 102/30, n=13), and 12,561 to 30,113 ml/min (mean/SD 18,730/5,464, n=18), respectively. Retrograde flow in the MPA was noted to begin at late-to mid-systole. Retrograde flow occurred first in the posterior part (15/19) or occurred from periphery (4/19). Thus hemodynamic parameters and velocity maps of MPA can be obtained by cine PCMRA. (author)

  5. Volumetric Arterial Wall Shear Stress Calculation Based on Cine Phase Contrast MRI

    NARCIS (Netherlands)

    Potters, Wouter V.; van Ooij, Pim; Marquering, Henk; VanBavel, Ed; Nederveen, Aart J.

    2015-01-01

    PurposeTo assess the accuracy and precision of a volumetric wall shear stress (WSS) calculation method applied to cine phase contrast magnetic resonance imaging (PC-MRI) data. Materials and MethodsVolumetric WSS vectors were calculated in software phantoms. WSS algorithm parameters were optimized

  6. Value of Cine-MRI sequences before and after injection in the diagnosis of acute myocarditis.

    Science.gov (United States)

    Zidi, Asma; Zairi, Ihsen; Mzoughi, Khadija; Zakhama, Lilia; Kamoun, Ikram; Ben Halima, Afef; Ridene, Imen

    2016-11-01

    Cardiovascular magnetic resonance (CMR) has become the examination of choice in case of suspicion of acute myocarditis. Late gadolinium enhancement (LGE) imaging is very important to establish this diagnosis. Cine MRI sequences are useful for the study of the myocardial contractility.   The purpose is to estimate the value of cine MRI sequences before and after injection for the diagnosis of acute myocarditis compared with late gadolinium enhanced sequences. We prospectively included 40 patients having a high suspicion of acute myocarditis and examined using a 1.5 Tesla CMR. Cine MRI sequences before and after injection were performed. The protocol also include  T2-weighted  short- tau-inversion-recovery (STIR T2) fast spin echo MRI and LGE imaging eight minutes after injection with visual adjustment of inversion time. Delayed enhancement was found among 23 patients. Fifteen patients (65 %) presented a spontaneous hyper signal detected visually on Cine MRI sequences before injection and 11 patients (48 %) on STIR T2. The hyper signal on Cine MRI sequences after injection of gadolinium was the same topography that the late raising at 23 patients. In addition, we highlighted a significant difference between this hyper signal before injection and the left ventricle ejection fraction (p=0.022) as well as with the telesystolic volume of the left ventricle (LV) indexed by the body mass (p=0.039). Our study suggests that Cine MRI sequences after injection are of equal performance in the diagnosis of acute myocarditis as the LGE sequences and its contibution is important when we want to shorten the examination or when inversion time isn't optimal.

  7. Functional MRI of the patellofemoral joint: comparison of ultrafast MRI, motion-triggered cine MRI and static MRI

    Energy Technology Data Exchange (ETDEWEB)

    Muhle, C. [Klinik fuer Radiologische Diagnostik, Univ. Kiel (Germany); Brossmann, J. [Klinik fuer Radiologische Diagnostik, Univ. Kiel (Germany); Melchert, U.H. [Klinik fuer Radiologische Diagnostik, Univ. Kiel (Germany); Schroeder, C. [Radiologische Abt., Universitaets-Kinderklinik, Christian-Albrechts-Universitaet, Kiel (Germany); Boer, R. de [Philips Medical Systems, Best (Netherlands); Spielmann, R.P. [Klinik fuer Radiologische Diagnostik, Univ. Kiel (Germany); Heller, M. [Klinik fuer Radiologische Diagnostik, Univ. Kiel (Germany)

    1995-12-31

    To evaluate the feasibility and usefulness of ultrafast MRI (u), patellar tracking from 30 of flexion to knee extension (0 ) was analysed and compared with motion-triggered cine MRI (m) and a static MRI technique (s). The different imaging methods were compared in respect of the patellofemoral relationship, the examination time and image quality. Eight healthy subjects and four patients (in total 18 joints) with patellar subluxation or luxation were examined. Significant differences between the static MRI series without quadriceps contraction and the functional MRI studies (motion-triggered cine MRI and ultrafast MRI) were found for the patellar tilt angle. In the dynamic joint studies there was no statistical difference of the regression coefficients between the motion-triggered cine MRI studies and the ultrafast MRI studies. The findings of the functional MRI studies compared with the static MRI images were significantly different for the lateralisation of the patella, expressed by the lateral patellar displacement and bisect offset. No significant differences in patellar lateralisation were found between motion-triggered cine MRI and ultrafast MRI. Ultrafast MRI was superior to motion-triggered cine MRI in terms of the reduction in