WorldWideScience

Sample records for ct number accuracy

  1. Accuracy of Dual-Energy Virtual Monochromatic CT Numbers: Comparison between the Single-Source Projection-Based and Dual-Source Image-Based Methods.

    Science.gov (United States)

    Ueguchi, Takashi; Ogihara, Ryota; Yamada, Sachiko

    2018-03-21

    To investigate the accuracy of dual-energy virtual monochromatic computed tomography (CT) numbers obtained by two typical hardware and software implementations: the single-source projection-based method and the dual-source image-based method. A phantom with different tissue equivalent inserts was scanned with both single-source and dual-source scanners. A fast kVp-switching feature was used on the single-source scanner, whereas a tin filter was used on the dual-source scanner. Virtual monochromatic CT images of the phantom at energy levels of 60, 100, and 140 keV were obtained by both projection-based (on the single-source scanner) and image-based (on the dual-source scanner) methods. The accuracy of virtual monochromatic CT numbers for all inserts was assessed by comparing measured values to their corresponding true values. Linear regression analysis was performed to evaluate the dependency of measured CT numbers on tissue attenuation, method, and their interaction. Root mean square values of systematic error over all inserts at 60, 100, and 140 keV were approximately 53, 21, and 29 Hounsfield unit (HU) with the single-source projection-based method, and 46, 7, and 6 HU with the dual-source image-based method, respectively. Linear regression analysis revealed that the interaction between the attenuation and the method had a statistically significant effect on the measured CT numbers at 100 and 140 keV. There were attenuation-, method-, and energy level-dependent systematic errors in the measured virtual monochromatic CT numbers. CT number reproducibility was comparable between the two scanners, and CT numbers had better accuracy with the dual-source image-based method at 100 and 140 keV. Copyright © 2018 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

  2. SU-F-J-205: Effect of Cone Beam Factor On Cone Beam CT Number Accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Yao, W; Hua, C; Farr, J; Brady, S; Merchant, T [St. Jude Children’s Research Hospital, Memphis, TN (United States)

    2016-06-15

    Purpose: To examine the suitability of a Catphan™ 700 phantom for image quality QA of a cone beam computed tomography (CBCT) system deployed for proton therapy. Methods: Catphan phantoms, particularly Catphan™ 504, are commonly used in image quality QA for CBCT. As a newer product, Catphan™ 700 offers more tissue equivalent inserts which may be useful for generating the electron density – CT number curve for CBCT based treatment planning. The sensitometry-and-geometry module used in Catphan™ 700 is located at the end of the phantom and after the resolution line pair module. In Catphan™ 504 the line pair module is located at the end of the phantom and after the sensitometry-and-geometry module. To investigate the effect of difference in location on CT number accuracy due to the cone beam factor, we scanned the Catphan™ 700 with the central plane of CBCT at the center of the phantom, line pair and sensitometry-andgeometry modules of the phantom, respectively. The protocol head and thorax scan modes were used. For each position, scans were repeated 4 times. Results: For the head scan mode, the standard deviation (SD) of the CT numbers of each insert under 4 repeated scans was up to 20 HU, 11 HU, and 11 HU, respectively, for the central plane of CBCT located at the center of the phantom, line pair, and sensitometry-and-geometry modules of the phantom. The mean of the SD was 9.9 HU, 5.7 HU, and 5.9 HU, respectively. For the thorax mode, the mean of the SD was 4.5 HU, 4.4 HU, and 4.4 HU, respectively. The assessment of image quality based on resolution and spatial linearity was not affected by imaging location changes. Conclusion: When the Catphan™ 700 was aligned to the center of imaging region, the CT number accuracy test may not meet expectations. We recommend reconfiguration of the modules.

  3. The effect of the cranial bone CT numbers on the brain CT numbers

    Energy Technology Data Exchange (ETDEWEB)

    Fukuda, Hitoshi; Kobayashi, Shotai; Koide, Hiromi; Yamaguchi, Shuhei; Okada, Kazunori; Shimote, Koichi; Tsunematsu, Tokugoro (Shimane Medical Univ., Izumo (Japan))

    1989-06-01

    The effects of the cranial size and the computed tomography (CT) numbers of the cranial bone on that of the brain were studied in 70 subjects, aged from 30 to 94 years. The subjects had no histories of cerebrovascular accidents and showed no abnormalities in the central nervous system upon physical examinations and a CT scan. We measured the average attenuation values (CT numbers) of each elliptical region (165 pixels, 0.39 cm{sup 2}) at the bilateral thalamus and at twelve areas of the deep white matter. Multiple regression analysis was used to assess the effects of age, cranial size, and cranial bone CT numbers on the brain CT numbers. The effect of the cranial bone CT numbers on the brain CT numbers was statistically significant. The brain CT numbers increased with the increase in the cranial bone CT numbers. There was, however, no significant correlation between brain CT numbers and cranial size. In measuring the brain CT numbers, it is desirable that consideration be given to the cranial bone CT numbers. (author).

  4. Transthoracic CT-guided biopsy with multiplanar reconstruction image improves diagnostic accuracy of solitary pulmonary nodules

    International Nuclear Information System (INIS)

    Ohno, Yoshiharu; Hatabu, Hiroto; Takenaka, Daisuke; Imai, Masatake; Ohbayashi, Chiho; Sugimura, Kazuro

    2004-01-01

    Objective: To evaluate the utility of multiplanar reconstruction (MPR) image for CT-guided biopsy and determine factors of influencing diagnostic accuracy and the pneumothorax rate. Materials and methods: 390 patients with 396 pulmonary nodules underwent transthoracic CT-guided aspiration biopsy (TNAB) and transthoracic CT-guided cutting needle core biopsy (TCNB) as follows: 250 solitary pulmonary nodules (SPNs) underwent conventional CT-guided biopsy (conventional method), 81 underwent CT-fluoroscopic biopsy (CT-fluoroscopic method) and 65 underwent conventional CT-guided biopsy in combination with MPR image (MPR method). Success rate, overall diagnostic accuracy, pneumothorax rate and total procedure time were compared in each method. Factors affecting diagnostic accuracy and pneumothorax rate of CT-guided biopsy were statistically evaluated. Results: Success rates (TNAB: 100.0%, TCNB: 100.0%) and overall diagnostic accuracies (TNAB: 96.9%, TCNB: 97.0%) of MPR were significantly higher than those using the conventional method (TNAB: 87.6 and 82.4%, TCNB: 86.3 and 81.3%) (P<0.05). Diagnostic accuracy were influenced by biopsy method, lesion size, and needle path length (P<0.05). Pneumothorax rate was influenced by pathological diagnostic method, lesion size, number of punctures and FEV1.0% (P<0.05). Conclusion: The use of MPR for CT-guided lung biopsy is useful for improving diagnostic accuracy with no significant increase in pneumothorax rate or total procedure time

  5. Comparison of CT numbers between cone-beam CT and multi-detector CT

    International Nuclear Information System (INIS)

    Kim, Dong Soo; Han, Won Jeong; Kim, Eun Kyung

    2010-01-01

    To compare the CT numbers on 3 cone-beam CT (CBCT) images with those on multi-detector CT (MDCT) image using CT phantom and to develop linear regressive equations using CT numbers to material density for all the CT scanner each. Mini CT phantom comprised of five 1 inch thick cylindrical models with 1.125 inches diameter of materials with different densities (polyethylene, polystyrene, plastic water, nylon and acrylic) was used. It was scanned in 3 CBCTs (i-CAT, Alphard VEGA, Implagraphy SC) and 1 MDCT (Somatom Emotion). The images were saved as DICOM format and CT numbers were measured using OnDemand 3D. CT numbers obtained from CBCTs and MDCT images were compared and linear regression analysis was performed for the density, ρ(g/cm 3 ), as the dependent variable in terms of the CT numbers obtained from CBCTs and MDCT images. CT numbers on i-CAT and Implagraphy CBCT images were smaller than those on Somatom Emotion MDCT image (p<0.05). Linear relationship on a range of materials used for this study were ρ=0.001 H+1.07 with R2 value of 0.999 for Somatom Emotion, ρ=0.002 H+1.09 with R2 value of 0.991 for Alphard VEGA, ρ=0.001 H+1.43 with R2 value of 0.980 for i-CAT and ρ=0.001 H+1.30 with R2 value of 0.975 for Implagraphy. CT numbers on i-CAT and Implagraphy CBCT images were not same as those on Somatom Emotion MDCT image. The linear regressive equations to determine the density from the CT numbers with very high correlation coefficient were obtained on three CBCT and MDCT scan.

  6. Comparison of CT numbers between cone-beam CT and multi-detector CT

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Dong Soo; Han, Won Jeong; Kim, Eun Kyung [Department of Oral and Maxillofacial Radiology, School of Dentistry, Dankook University, Cheonan (Korea, Republic of)

    2010-06-15

    To compare the CT numbers on 3 cone-beam CT (CBCT) images with those on multi-detector CT (MDCT) image using CT phantom and to develop linear regressive equations using CT numbers to material density for all the CT scanner each. Mini CT phantom comprised of five 1 inch thick cylindrical models with 1.125 inches diameter of materials with different densities (polyethylene, polystyrene, plastic water, nylon and acrylic) was used. It was scanned in 3 CBCTs (i-CAT, Alphard VEGA, Implagraphy SC) and 1 MDCT (Somatom Emotion). The images were saved as DICOM format and CT numbers were measured using OnDemand 3D. CT numbers obtained from CBCTs and MDCT images were compared and linear regression analysis was performed for the density, {rho}(g/cm{sup 3}), as the dependent variable in terms of the CT numbers obtained from CBCTs and MDCT images. CT numbers on i-CAT and Implagraphy CBCT images were smaller than those on Somatom Emotion MDCT image (p<0.05). Linear relationship on a range of materials used for this study were {rho}=0.001 H+1.07 with R2 value of 0.999 for Somatom Emotion, {rho}=0.002 H+1.09 with R2 value of 0.991 for Alphard VEGA, {rho}=0.001 H+1.43 with R2 value of 0.980 for i-CAT and {rho}=0.001 H+1.30 with R2 value of 0.975 for Implagraphy. CT numbers on i-CAT and Implagraphy CBCT images were not same as those on Somatom Emotion MDCT image. The linear regressive equations to determine the density from the CT numbers with very high correlation coefficient were obtained on three CBCT and MDCT scan.

  7. Accuracy of radiographer reporting of paediatric brain CT

    International Nuclear Information System (INIS)

    Brandt, Andrew; Louw, Brand; Dekker, Gerrit; Andronikou, Savvas; Wieselthaler, Nicki; Kilborn, Tracy; Bertelsman, Jessica; Dreyer, Catherine

    2007-01-01

    Radiographer reporting has been studied for plain films and for ultrasonography, but not in paediatric brain CT in the emergency setting. To study the accuracy of radiographer reporting in paediatric brain CT. We prospectively collected 100 paediatric brain CT examinations. Films were read from hard copies using a prescribed tick sheet. Radiographers with 12 years' and 3 years' experience, respectively, were blinded to the history and were not trained in diagnostic film interpretation. The radiographers' results were compared with those of a consultant radiologist. Three categories were defined: abnormal scans, significant abnormalities and insignificant abnormalities. Both radiographers had an accuracy of 89.5% in reading a scan correctly as abnormal, and radiographer 1 had a sensitivity of 87.8% and radiographer 2 a sensitivity of 96%. Radiographer 1 had an accuracy in detecting a significant abnormality of 75% and radiographer 2 an accuracy of 48.6%, and the sensitivities for this category were 61.6% and 52.9%, respectively. Results for detecting the insignificant abnormalities were poorer. Selected radiographers could play an effective screening role, but lacking the sensitivity required for detecting significant abnormality, they could not be the final diagnostician. We recommend that the study be repeated after both radiographers have received formal training in interpretation of paediatric brain CT. (orig.)

  8. CT number of the fatty liver

    International Nuclear Information System (INIS)

    Maeda, Hiroko; Kawai, Takeshi; Kanasaki, Yoshiki; Akagi, Hiroaki

    1981-01-01

    This report is studied on CT number and CT images of the eight cases with fatty liver. Five of these cases showed the reversal of densities of the liver and vessels. In these cases, the diagnoses of the fatty liver were easible. In other cases, the diagnoses were possible only by comparison of the CT number of the liver and spleen because the CT number of normal liver were higher than those of the spleen. In the results which we examined the correlation of the CT number and specific gravities of the blood, normal saline, distilled water, mayonnaise, eatable iol, ethyl alcohol and lard, we observed the linear relationship between CT number and specific gravities. And so, we think that the diagnosis of the fatty liver and the degree of fatty infiltration can be guessed by the CT number of the liver and spleen. (author)

  9. Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions

    Science.gov (United States)

    Schneider, Wilfried; Bortfeld, Thomas; Schlegel, Wolfgang

    2000-02-01

    We describe a new method to convert CT numbers into mass density and elemental weights of tissues required as input for dose calculations with Monte Carlo codes such as EGS4. As a first step, we calculate the CT numbers for 71 human tissues. To reduce the effort for the necessary fits of the CT numbers to mass density and elemental weights, we establish four sections on the CT number scale, each confined by selected tissues. Within each section, the mass density and elemental weights of the selected tissues are interpolated. For this purpose, functional relationships between the CT number and each of the tissue parameters, valid for media which are composed of only two components in varying proportions, are derived. Compared with conventional data fits, no loss of accuracy is accepted when using the interpolation functions. Assuming plausible values for the deviations of calculated and measured CT numbers, the mass density can be determined with an accuracy better than 0.04 g cm-3 . The weights of phosphorus and calcium can be determined with maximum uncertainties of 1 or 2.3 percentage points (pp) respectively. Similar values can be achieved for hydrogen (0.8 pp) and nitrogen (3 pp). For carbon and oxygen weights, errors up to 14 pp can occur. The influence of the elemental weights on the results of Monte Carlo dose calculations is investigated and discussed.

  10. Accuracy of CT-guided biopsies in 158 patients with thoracic spinal lesions

    Energy Technology Data Exchange (ETDEWEB)

    Hao, D.J.; He, B.R.; Liu, T.J.; Zhao, Q.P. (Dept. of Spinal Surgery, Xian Red Cross Hospital, Xian Shaanxi (China)), email: zqpddn1@gmail.com; Sun, H.H. (Dept. of Orthopaedic, Tangdu Hospital, Fourth Military Medical Univ., Xian Shaanxi (China)); Jiang, Y.H. (Dept. of Radiology, Xian Red Cross Hospital, Xian Shaanxi (China))

    2011-11-15

    Background. Inconsistent accuracies of CT-guided thoracic spinal biopsies have been reported in previous studies. Purpose. To determine the accuracy of CT-guided thoracic spinal biopsy, to compare the results with those previously reported, and to determine if there are any factors that influence the accuracy of CT-guided thoracic spinal biopsy. Material and Methods. In total, 158 consecutive CT-guided percutaneous thoracic spine procedures (performed at the Dept. of Spinal Surgery, Xian Red Cross Hospital between April 2000 and July 2010) were reviewed. The 158 lesions were categorized by location and radiographic features. Pathological and clinical follow-up were used to determine accuracy. Results. The diagnostic accuracy of CT-guided thoracic spinal biopsy was 90.5% overall. Biopsy of metastatic bone disease (98.2%) was significantly more accurate than biopsies of primary tumors (80.9%) and of hematological malignancies (47.0%) (P < 0.05 and P < 0.005, respectively). The diagnostic accuracy of CT-guided thoracic spinal biopsy was significantly higher for the lower thoracic spine (97.6%) than for the middle (90.0%) or upper thoracic spine (80.4%) (P < 0.05 and P < 0.025, respectively). The diagnostic accuracy was significantly higher for lytic lesions (96.4%) than for sclerotic lesions (81.3%) (P < 0.010). The accuracy of biopsies performed using the transpedicular approach (91.0%) was not significantly different from that of biopsies performed using posterolateral approaches (91.5%) (0.25 < P < 0.5). Conclusion. Percutaneous CT-guided thoracic spinal biopsy is a viable alternative to open surgical biopsy. The diagnostic accuracy was not affected by any of the variables except for lesion level, histology, and radiographic features

  11. Accuracy of CT-guided biopsies in 158 patients with thoracic spinal lesions

    International Nuclear Information System (INIS)

    Hao, D.J.; He, B.R.; Liu, T.J.; Zhao, Q.P.; Sun, H.H.; Jiang, Y.H.

    2011-01-01

    Background. Inconsistent accuracies of CT-guided thoracic spinal biopsies have been reported in previous studies. Purpose. To determine the accuracy of CT-guided thoracic spinal biopsy, to compare the results with those previously reported, and to determine if there are any factors that influence the accuracy of CT-guided thoracic spinal biopsy. Material and Methods. In total, 158 consecutive CT-guided percutaneous thoracic spine procedures (performed at the Dept. of Spinal Surgery, Xian Red Cross Hospital between April 2000 and July 2010) were reviewed. The 158 lesions were categorized by location and radiographic features. Pathological and clinical follow-up were used to determine accuracy. Results. The diagnostic accuracy of CT-guided thoracic spinal biopsy was 90.5% overall. Biopsy of metastatic bone disease (98.2%) was significantly more accurate than biopsies of primary tumors (80.9%) and of hematological malignancies (47.0%) (P < 0.05 and P < 0.005, respectively). The diagnostic accuracy of CT-guided thoracic spinal biopsy was significantly higher for the lower thoracic spine (97.6%) than for the middle (90.0%) or upper thoracic spine (80.4%) (P < 0.05 and P < 0.025, respectively). The diagnostic accuracy was significantly higher for lytic lesions (96.4%) than for sclerotic lesions (81.3%) (P < 0.010). The accuracy of biopsies performed using the transpedicular approach (91.0%) was not significantly different from that of biopsies performed using posterolateral approaches (91.5%) (0.25 < P < 0.5). Conclusion. Percutaneous CT-guided thoracic spinal biopsy is a viable alternative to open surgical biopsy. The diagnostic accuracy was not affected by any of the variables except for lesion level, histology, and radiographic features

  12. Detection of relevant colonic neoplasms with PET/CT: promising accuracy with minimal CT dose and a standardised PET cut-off

    Energy Technology Data Exchange (ETDEWEB)

    Luboldt, Wolfgang [Multiorgan Screening Foundation, Frankfurt (Germany); University Hospital Frankfurt, Department of Radiology, Frankfurt am Main (Germany); University Hospital Dresden, Clinic and Policlinic of Nuclear Medicine, Dresden (Germany); Volker, Teresa; Zoephel, Klaus; Kotzerke, Joerg [University Hospital Dresden, Clinic and Policlinic of Nuclear Medicine, Dresden (Germany); Wiedemann, Baerbel [University Hospital Dresden, Institute of Medical Informatics and Biometrics, Dresden (Germany); Wehrmann, Ursula [University Hospital Dresden, Clinic and Policlinic of Surgery, Dresden (Germany); Koch, Arne; Abolmaali, Nasreddin [University Hospital Dresden, Oncoray, Dresden (Germany); Toussaint, Todd; Luboldt, Hans-Joachim [Multiorgan Screening Foundation, Frankfurt (Germany); Middendorp, Markus; Gruenwald, Frank [University Hospital Frankfurt, Department of Nuclear Medicine, Frankfurt (Germany); Aust, Daniela [University Hospital Dresden, Department of Pathology, Dresden (Germany); Vogl, Thomas J. [University Hospital Frankfurt, Department of Radiology, Frankfurt am Main (Germany)

    2010-09-15

    To determine the performance of FDG-PET/CT in the detection of relevant colorectal neoplasms (adenomas {>=}10 mm, with high-grade dysplasia, cancer) in relation to CT dose and contrast administration and to find a PET cut-off. 84 patients, who underwent PET/CT and colonoscopy (n=79)/sigmoidoscopy (n=5) for (79 x 6+5 x 2)=484 colonic segments, were included in a retrospective study. The accuracy of low-dose PET/CT in detecting mass-positive segments was evaluated by ROC analysis by two blinded independent reviewers relative to contrast-enhanced PET/CT. On a per-lesion basis characteristic PET values were tested as cut-offs. Low-dose PET/CT and contrast-enhanced PET/CT provide similar accuracies (area under the curve for the average ROC ratings 0.925 vs. 0.929, respectively). PET demonstrated all carcinomas (n=23) and 83% (30/36) of relevant adenomas. In all carcinomas and adenomas with high-grade dysplasia (n=10) the SUV{sub max} was {>=}5. This cut-off resulted in a better per-segment sensitivity and negative predictive value (NPV) than the average PET/CT reviews (sensitivity: 89% vs. 82%; NPV: 99% vs. 98%). All other tested cut-offs were inferior to the SUV{sub max}. FDG-PET/CT provides promising accuracy for colorectal mass detection. Low dose and lack of iodine contrast in the CT component do not impact the accuracy. The PET cut-off SUV{sub max}{>=} 5 improves the accuracy. (orig.)

  13. Detection of relevant colonic neoplasms with PET/CT: promising accuracy with minimal CT dose and a standardised PET cut-off

    International Nuclear Information System (INIS)

    Luboldt, Wolfgang; Volker, Teresa; Zoephel, Klaus; Kotzerke, Joerg; Wiedemann, Baerbel; Wehrmann, Ursula; Koch, Arne; Abolmaali, Nasreddin; Toussaint, Todd; Luboldt, Hans-Joachim; Middendorp, Markus; Gruenwald, Frank; Aust, Daniela; Vogl, Thomas J.

    2010-01-01

    To determine the performance of FDG-PET/CT in the detection of relevant colorectal neoplasms (adenomas ≥10 mm, with high-grade dysplasia, cancer) in relation to CT dose and contrast administration and to find a PET cut-off. 84 patients, who underwent PET/CT and colonoscopy (n=79)/sigmoidoscopy (n=5) for (79 x 6+5 x 2)=484 colonic segments, were included in a retrospective study. The accuracy of low-dose PET/CT in detecting mass-positive segments was evaluated by ROC analysis by two blinded independent reviewers relative to contrast-enhanced PET/CT. On a per-lesion basis characteristic PET values were tested as cut-offs. Low-dose PET/CT and contrast-enhanced PET/CT provide similar accuracies (area under the curve for the average ROC ratings 0.925 vs. 0.929, respectively). PET demonstrated all carcinomas (n=23) and 83% (30/36) of relevant adenomas. In all carcinomas and adenomas with high-grade dysplasia (n=10) the SUV max was ≥5. This cut-off resulted in a better per-segment sensitivity and negative predictive value (NPV) than the average PET/CT reviews (sensitivity: 89% vs. 82%; NPV: 99% vs. 98%). All other tested cut-offs were inferior to the SUV max . FDG-PET/CT provides promising accuracy for colorectal mass detection. Low dose and lack of iodine contrast in the CT component do not impact the accuracy. The PET cut-off SUV max ≥ 5 improves the accuracy. (orig.)

  14. CT Angiography for Living Kidney Donors: Accuracy, Cause of Misinterpretation and Prevalence of Variation

    Energy Technology Data Exchange (ETDEWEB)

    Chai, Jee Won; Lee, Whal; Yin, Yong Hu; Jae, Hwan Jun; Chung, Jin Wook; Park, Jae Hyung [Seoul National University Hospital, Seoul (Korea, Republic of); Kim, Hyeon Hoe [Seoul National University College of Medicine, Seoul (Korea, Republic of)

    2008-08-15

    To determine the accuracy of the use of multi-detector row CT (MDCT) to predict vascular anatomy in living kidney donors and to reveal the prevalence of vascular variations in a Korean population. A total of 153 living kidney donors that had undergone preoperative CT and nephrectomy, either with open or laparoscopic surgery, were selected retrospectively. The initial CT results were compared with the surgical findings and repeated review sessions of CT scans were performed to determine the causes of mismatches in discordant cases. The accuracy of CT angiography was 95% to predict the number of renal vessels. Four arteries and two veins were missed during the initial CT interpretation due to perception errors (for two arteries and two veins) and technical limitations (two arteries). The prevalence of multiple renal arteries and veins, early branching of a renal artery and late confluence of a renal vein were 31%, 5%, 12%, 17%, respectively. The circumaortic renal vein and the bilateral inferior vena cava were found in two cases each (1.3%). One case (0.7%) each of a retroaortic renal vein and a supradiaphragmatic originated renal artery were found. MDCT provides a reliable method to evaluate the vascular anatomy and variations of living kidney donors.

  15. CT Angiography for Living Kidney Donors: Accuracy, Cause of Misinterpretation and Prevalence of Variation

    International Nuclear Information System (INIS)

    Chai, Jee Won; Lee, Whal; Yin, Yong Hu; Jae, Hwan Jun; Chung, Jin Wook; Park, Jae Hyung; Kim, Hyeon Hoe

    2008-01-01

    To determine the accuracy of the use of multi-detector row CT (MDCT) to predict vascular anatomy in living kidney donors and to reveal the prevalence of vascular variations in a Korean population. A total of 153 living kidney donors that had undergone preoperative CT and nephrectomy, either with open or laparoscopic surgery, were selected retrospectively. The initial CT results were compared with the surgical findings and repeated review sessions of CT scans were performed to determine the causes of mismatches in discordant cases. The accuracy of CT angiography was 95% to predict the number of renal vessels. Four arteries and two veins were missed during the initial CT interpretation due to perception errors (for two arteries and two veins) and technical limitations (two arteries). The prevalence of multiple renal arteries and veins, early branching of a renal artery and late confluence of a renal vein were 31%, 5%, 12%, 17%, respectively. The circumaortic renal vein and the bilateral inferior vena cava were found in two cases each (1.3%). One case (0.7%) each of a retroaortic renal vein and a supradiaphragmatic originated renal artery were found. MDCT provides a reliable method to evaluate the vascular anatomy and variations of living kidney donors

  16. A simulation study on proton computed tomography (CT) stopping power accuracy using dual energy CT scans as benchmark

    DEFF Research Database (Denmark)

    Hansen, David Christoffer; Seco, Joao; Sørensen, Thomas Sangild

    2015-01-01

    Background. Accurate stopping power estimation is crucial for treatment planning in proton therapy, and the uncertainties in stopping power are currently the largest contributor to the employed dose margins. Dual energy x-ray computed tomography (CT) (clinically available) and proton CT (in...... development) have both been proposed as methods for obtaining patient stopping power maps. The purpose of this work was to assess the accuracy of proton CT using dual energy CT scans of phantoms to establish reference accuracy levels. Material and methods. A CT calibration phantom and an abdomen cross section...... phantom containing inserts were scanned with dual energy and single energy CT with a state-of-the-art dual energy CT scanner. Proton CT scans were simulated using Monte Carlo methods. The simulations followed the setup used in current prototype proton CT scanners and included realistic modeling...

  17. CT numbers of liver and spleen in normal children

    International Nuclear Information System (INIS)

    Kim, Young Kim

    2002-01-01

    To determine the mean liver CT numbers, and differences between liver and spleen, and liver and back muscle CT numbers in normal children, and to correlate the findings with sex and age. One hundred and five normal children aged 2-14 years underwent pre-contrast CT scanning. Mean CT numbers of the liver, spleen, and back muscles were calculated, as well as the differences in CT numbers between the liver and spleen (liver-spleen CT numbers), and between the liver-back muscle CT numbers were 70.22±6.51 HU, 53.28±3.57 HU, 17.13±6.57 HU, and 11.88±5.94 HU, respectively. Mean liver CT numbers and the difference between liver and back muscle CT numbers were not different by age. By sex, all the CT numbers did not vary according to age. The sex of a subject did not affect the CT number. The children's mean liver CT number was 70.22±6.51 HU and the difference between liver and spleen CT numbers was 17.13±6.57 HU. Younger children had higher liver CT and liver-spleen CT numbers than older children. No CT numbers varied according to sex

  18. The accuracy of after-hour registrar computed tomography (CT ...

    African Journals Online (AJOL)

    2014-05-16

    May 16, 2014 ... tomography (CT) reporting in a South African tertiary teaching hospital. Authors: ... The purpose of this study was to determine the accuracy of after-hour registrar CT reporting, to identify possible factors that may .... The authors declare that they have no financial or personal relationship(s) that may have ...

  19. Reconstructing the CT number array from gray-level images and its application in PACS

    Science.gov (United States)

    Chen, Xu; Zhuang, Tian-ge; Wu, Wei

    2001-08-01

    Although DICOM compliant computed tomography has been prevailing in medical fields nowadays, there are some incompliant ones, from which we could hardly get the raw data and make an apropos interpretation due to the proprietary image format. Under such condition, one usually uses frame grabbers to capture CT images, the results of which could not be freely adjusted by radiologists as the original CT number array could. To alleviate the inflexibility, a new method is presented in this paper to reconstruct the array of CT number from several gray-level images acquired under different window settings. Its feasibility is investigated and a few tips are put forward to correct the errors caused respectively by 'Border Effect' and some hardware problems. The accuracy analysis proves it a good substitution for original CT number array acquisition. And this method has already been successfully used in our newly developing PACS and accepted by the radiologists in clinical use.

  20. Energy Dependence of Measured CT Numbers on Substituted Materials Used for CT Number Calibration of Radiotherapy Treatment Planning Systems.

    Directory of Open Access Journals (Sweden)

    Reza Mahmoudi

    Full Text Available For accurate dose calculations, it is necessary to provide a correct relationship between the CT numbers and electron density in radiotherapy treatment planning systems (TPSs. The purpose of this study was to investigate the energy dependence of measured CT numbers on substituted materials used for CT number calibration of radiotherapy TPSs and the resulting errors in the treatment planning calculation doses.In this study, we designed a cylindrical water phantom with different materials used as tissue equivalent materials for the simulation of tissues and obtaining the related CT numbers. For evaluating the effect of CT number variations of substituted materials due to energy changing of scanner (kVp on the dose calculation of TPS, the slices of the scanned phantom at three kVp's were imported into the desired TPSs (MIRS and CorePLAN. Dose calculations were performed on two TPSs.The mean absolute percentage differences between the CT numbers of CT scanner and two treatment planning systems for all the samples were 3.22%±2.57% for CorePLAN and 2.88%±2.11% for MIRS. It was also found that the maximum absolute percentage difference between all of the calculated doses from each photon beam of linac (6 and 15 MV at three kVp's was less than 1.2%.The present study revealed that, for the materials with effective low atomic number, the mean CT number increased with increasing energy, which was opposite for the materials with an effective high atomic number. We concluded that the tissue substitute materials had a different behavior in the energy ranges from 80 to 130 kVp. So, it is necessary to consider the energy dependence of the substitute materials used for the measurement or calibration of CT number for radiotherapy treatment planning systems.

  1. Stereotactic imaging for radiotherapy: accuracy of CT, MRI, PET and SPECT

    International Nuclear Information System (INIS)

    Karger, Christian P; Hipp, Peter; Henze, Marcus; Echner, Gernot; Hoess, Angelika; Schad, Lothar; Hartmann, Guenther H

    2003-01-01

    CT, MRI, PET and SPECT provide complementary information for treatment planning in stereotactic radiotherapy. Stereotactic correlation of these images requires commissioning tests to confirm the localization accuracy of each modality. A phantom was developed to measure the accuracy of stereotactic localization for CT, MRI, PET and SPECT in the head and neck region. To this end, the stereotactically measured coordinates of structures within the phantom were compared with their mechanically defined coordinates. For MRI, PET and SPECT, measurements were performed using two different devices. For MRI, T1- and T2-weighted imaging sequences were applied. For each measurement, the mean radial deviation in space between the stereotactically measured and mechanically defined position of target points was determined. For CT, the mean radial deviation was 0.4 ± 0.2 mm. For MRI, the mean deviations ranged between 0.7 ± 0.2 mm and 1.4 ± 0.5 mm, depending on the MRI device and the imaging sequence. For PET, mean deviations of 1.1 ± 0.5 mm and 2.4 ± 0.3 mm were obtained. The mean deviations for SPECT were 1.6 ± 0.5 mm and 2.0 ± 0.6 mm. The phantom is well suited to determine the accuracy of stereotactic localization with CT, MRI, PET and SPECT in the head and neck region. The obtained accuracy is well below the physical resolution for CT, PET and SPECT, and of comparable magnitude for MRI. Since the localization accuracy may be device dependent, results obtained at one device cannot be generalized to others

  2. SU-F-T-441: Dose Calculation Accuracy in CT Images Reconstructed with Artifact Reduction Algorithm

    Energy Technology Data Exchange (ETDEWEB)

    Ng, C; Chan, S; Lee, F; Ngan, R [Queen Elizabeth Hospital (Hong Kong); Lee, V [University of Hong Kong, Hong Kong, HK (Hong Kong)

    2016-06-15

    Purpose: Accuracy of radiotherapy dose calculation in patients with surgical implants is complicated by two factors. First is the accuracy of CT number, second is the dose calculation accuracy. We compared measured dose with dose calculated on CT images reconstructed with FBP and an artifact reduction algorithm (OMAR, Philips) for a phantom with high density inserts. Dose calculation were done with Varian AAA and AcurosXB. Methods: A phantom was constructed with solid water in which 2 titanium or stainless steel rods could be inserted. The phantom was scanned with the Philips Brillance Big Bore CT. Image reconstruction was done with FBP and OMAR. Two 6 MV single field photon plans were constructed for each phantom. Radiochromic films were placed at different locations to measure the dose deposited. One plan has normal incidence on the titanium/steel rods. In the second plan, the beam is at almost glancing incidence on the metal rods. Measurements were then compared with dose calculated with AAA and AcurosXB. Results: The use of OMAR images slightly improved the dose calculation accuracy. The agreement between measured and calculated dose was best with AXB and image reconstructed with OMAR. Dose calculated on titanium phantom has better agreement with measurement. Large discrepancies were seen at points directly above and below the high density inserts. Both AAA and AXB underestimated the dose directly above the metal surface, while overestimated the dose below the metal surface. Doses measured downstream of metal were all within 3% of calculated values. Conclusion: When doing treatment planning for patients with metal implants, care must be taken to acquire correct CT images to improve dose calculation accuracy. Moreover, great discrepancies in measured and calculated dose were observed at metal/tissue interface. Care must be taken in estimating the dose in critical structures that come into contact with metals.

  3. Analysis of diagnositc accuracy with CT-guided percutaneous transthoracic lung biopsy for pulmonary lesions

    International Nuclear Information System (INIS)

    Wang Haiyan; Zhuang Yiping; Zhang Jin; Feng Yong; Zhang Xiao; Xu Lin

    2007-01-01

    Objective: To evaluate the factors which may affect the diagnostic accmracy of CT-guided percutaneous transthoracic lung biopsy for pulmonary lesions. Methods: CT-guided percutaneous transthoracic lung biopsy was performed in 264 patients, all of which were proved by surgical pathology or clinical follow-up. A logistic regression analysis was used to investigate the relationship between the location and the size of the lesions and insertion depth of biopsy needles and the diagnostic accuracy of CT-guided percutaneous transthomcic lung biopsy. The complication rate including pneumatothorax and bleeding rate were record. Results: The diagnostic accuracy was 87.5%. Both the pneumothorax and the bleeding of lesion were 1.9%. The size of lesion was suggested by Logistic regression analysis had a significant influence on the diagnostic accuracy (Wald=12.891, P=0.00034, OR=0.96). Conclusion: CT-guided pereumneous transthoracic lung biopsy has a high diagnostic accuracy for pulmonary lesions. The size of the lesions had significant effect on the diagnostic accuracy. (authors)

  4. Integration of PET-CT and cone-beam CT for image-guided radiotherapy with high image quality and registration accuracy

    Science.gov (United States)

    Wu, T.-H.; Liang, C.-H.; Wu, J.-K.; Lien, C.-Y.; Yang, B.-H.; Huang, Y.-H.; Lee, J. J. S.

    2009-07-01

    Hybrid positron emission tomography-computed tomography (PET-CT) system enhances better differentiation of tissue uptake of 18F-fluorodeoxyglucose (18F-FDG) and provides much more diagnostic value in the non-small-cell lung cancer and nasopharyngeal carcinoma (NPC). In PET-CT, high quality CT images not only offer diagnostic value on anatomic delineation of the tissues but also shorten the acquisition time for attenuation correction (AC) compared with PET-alone imaging. The linear accelerators equipped with the X-ray cone-beam computed tomography (CBCT) imaging system for image-guided radiotherapy (IGRT) provides excellent verification on position setup error. The purposes of our study were to optimize the CT acquisition protocols of PET-CT and to integrate the PET-CT and CBCT for IGRT. The CT imaging parameters were modified in PET-CT for increasing the image quality in order to enhance the diagnostic value on tumour delineation. Reproducibility and registration accuracy via bone co-registration algorithm between the PET-CT and CBCT were evaluated by using a head phantom to simulate a head and neck treatment condition. Dose measurement in computed tomography dose index (CTDI) was also estimated. Optimization of the CT acquisition protocols of PET-CT was feasible in this study. Co-registration accuracy between CBCT and PET-CT on axial and helical modes was in the range of 1.06 to 2.08 and 0.99 to 2.05 mm, respectively. In our result, it revealed that the accuracy of the co-registration with CBCT on helical mode was more accurate than that on axial mode. Radiation doses in CTDI were 4.76 to 18.5 mGy and 4.83 to 18.79 mGy on axial and helical modes, respectively. Registration between PET-CT and CBCT is a state-of-the-art registration technology which could provide much information on diagnosis and accurate tumour contouring on radiotherapy while implementing radiotherapy procedures. This novelty technology of PET-CT and cone-beam CT integration for IGRT may have a

  5. Integration of PET-CT and cone-beam CT for image-guided radiotherapy with high image quality and registration accuracy

    International Nuclear Information System (INIS)

    Wu, T-H; Liang, C-H; Wu, J-K; Lien, C-Y; Yang, B-H; Lee, J J S; Huang, Y-H

    2009-01-01

    Hybrid positron emission tomography-computed tomography (PET-CT) system enhances better differentiation of tissue uptake of 18 F-fluorodeoxyglucose ( 18 F-FDG) and provides much more diagnostic value in the non-small-cell lung cancer and nasopharyngeal carcinoma (NPC). In PET-CT, high quality CT images not only offer diagnostic value on anatomic delineation of the tissues but also shorten the acquisition time for attenuation correction (AC) compared with PET-alone imaging. The linear accelerators equipped with the X-ray cone-beam computed tomography (CBCT) imaging system for image-guided radiotherapy (IGRT) provides excellent verification on position setup error. The purposes of our study were to optimize the CT acquisition protocols of PET-CT and to integrate the PET-CT and CBCT for IGRT. The CT imaging parameters were modified in PET-CT for increasing the image quality in order to enhance the diagnostic value on tumour delineation. Reproducibility and registration accuracy via bone co-registration algorithm between the PET-CT and CBCT were evaluated by using a head phantom to simulate a head and neck treatment condition. Dose measurement in computed tomography dose index (CTDI) was also estimated. Optimization of the CT acquisition protocols of PET-CT was feasible in this study. Co-registration accuracy between CBCT and PET-CT on axial and helical modes was in the range of 1.06 to 2.08 and 0.99 to 2.05 mm, respectively. In our result, it revealed that the accuracy of the co-registration with CBCT on helical mode was more accurate than that on axial mode. Radiation doses in CTDI were 4.76 to 18.5 mGy and 4.83 to 18.79 mGy on axial and helical modes, respectively. Registration between PET-CT and CBCT is a state-of-the-art registration technology which could provide much information on diagnosis and accurate tumour contouring on radiotherapy while implementing radiotherapy procedures. This novelty technology of PET-CT and cone-beam CT integration for IGRT may have a

  6. A simulation study on proton computed tomography (CT) stopping power accuracy using dual energy CT scans as benchmark.

    Science.gov (United States)

    Hansen, David C; Seco, Joao; Sørensen, Thomas Sangild; Petersen, Jørgen Breede Baltzer; Wildberger, Joachim E; Verhaegen, Frank; Landry, Guillaume

    2015-01-01

    Accurate stopping power estimation is crucial for treatment planning in proton therapy, and the uncertainties in stopping power are currently the largest contributor to the employed dose margins. Dual energy x-ray computed tomography (CT) (clinically available) and proton CT (in development) have both been proposed as methods for obtaining patient stopping power maps. The purpose of this work was to assess the accuracy of proton CT using dual energy CT scans of phantoms to establish reference accuracy levels. A CT calibration phantom and an abdomen cross section phantom containing inserts were scanned with dual energy and single energy CT with a state-of-the-art dual energy CT scanner. Proton CT scans were simulated using Monte Carlo methods. The simulations followed the setup used in current prototype proton CT scanners and included realistic modeling of detectors and the corresponding noise characteristics. Stopping power maps were calculated for all three scans, and compared with the ground truth stopping power from the phantoms. Proton CT gave slightly better stopping power estimates than the dual energy CT method, with root mean square errors of 0.2% and 0.5% (for each phantom) compared to 0.5% and 0.9%. Single energy CT root mean square errors were 2.7% and 1.6%. Maximal errors for proton, dual energy and single energy CT were 0.51%, 1.7% and 7.4%, respectively. Better stopping power estimates could significantly reduce the range errors in proton therapy, but requires a large improvement in current methods which may be achievable with proton CT.

  7. Unilateral Hemithorax Opacification on Chest Radiograph : Comparison of Diagnostic Accuracy of Chest Ultrasonography with CT

    Energy Technology Data Exchange (ETDEWEB)

    Namkung, Sook; Lee, Kyung Soo; Kwon, O Jung [Samsung Medical Center, Seoul (Korea, Republic of); Chung, Seung Eun [Sangkye Paek Hospital, Seoul (Korea, Republic of)

    1996-12-15

    To compare the diagnostic accuracy of chest ultrasonography (US) with computed tomography (CT) inpatients with opacification more than one third of unilateral hemithorax on chest radiograph (CXR). Chest US and CT scans from 41 consecutive patients with opacification more than one third of unilateral hemithorax on CXR were prospectively evaluated by two independent radiologists. Each radiologist recorded 1) the nature of pleural effusion (transudate vs. exudate), 2) presence or absence of pulmonary lesion, 3) the characteristic of pulmonary lesion (consolidation or atelectasis and tumor), and 4) presence of solid pleural tumor. The diagnostic accuracy of chest US was compared with CT scan in patients with pleural, pulmonary or other disease. In 32 patients with pleural effusion, differentiation between transudate and exudate was feasible in 27 (84%) patients with US and 26 (81%) patients with CT. In 32 patients with pulmonary and other pleural diseases, sensitivity, specificity and accuracy of US in lesion detection were 86%, 75% and 83% respectively when CT was regarded as a diagnostic gold standard. The diagnostic accuracy of chest US is comparable to CT in patients with hemithorax opacification on CXR

  8. Unilateral Hemithorax Opacification on Chest Radiograph : Comparison of Diagnostic Accuracy of Chest Ultrasonography with CT

    International Nuclear Information System (INIS)

    Namkung, Sook; Lee, Kyung Soo; Kwon, O Jung; Chung, Seung Eun

    1996-01-01

    To compare the diagnostic accuracy of chest ultrasonography (US) with computed tomography (CT) inpatients with opacification more than one third of unilateral hemithorax on chest radiograph (CXR). Chest US and CT scans from 41 consecutive patients with opacification more than one third of unilateral hemithorax on CXR were prospectively evaluated by two independent radiologists. Each radiologist recorded 1) the nature of pleural effusion (transudate vs. exudate), 2) presence or absence of pulmonary lesion, 3) the characteristic of pulmonary lesion (consolidation or atelectasis and tumor), and 4) presence of solid pleural tumor. The diagnostic accuracy of chest US was compared with CT scan in patients with pleural, pulmonary or other disease. In 32 patients with pleural effusion, differentiation between transudate and exudate was feasible in 27 (84%) patients with US and 26 (81%) patients with CT. In 32 patients with pulmonary and other pleural diseases, sensitivity, specificity and accuracy of US in lesion detection were 86%, 75% and 83% respectively when CT was regarded as a diagnostic gold standard. The diagnostic accuracy of chest US is comparable to CT in patients with hemithorax opacification on CXR

  9. A comprehensive study on decreasing the kilovoltage cone-beam CT dose by reducing the projection number.

    Science.gov (United States)

    Lu, Bo; Lu, Haibin; Palta, Jatinder

    2010-05-12

    The objective of this study was to evaluate the effect of kilovoltage cone-beam computed tomography (CBCT) on registration accuracy and image qualities with a reduced number of planar projections used in volumetric imaging reconstruction. The ultimate goal is to evaluate the possibility of reducing the patient dose while maintaining registration accuracy under different projection-number schemes for various clinical sites. An Elekta Synergy Linear accelerator with an onboard CBCT system was used in this study. The quality of the Elekta XVI cone-beam three-dimensional volumetric images reconstructed with a decreasing number of projections was quantitatively evaluated by a Catphan phantom. Subsequently, we tested the registration accuracy of imaging data sets on three rigid anthropomorphic phantoms and three real patient sites under the reduced projection-number (as low as 1/6th) reconstruction of CBCT data with different rectilinear shifts and rota-tions. CBCT scan results of the Catphan phantom indicated the CBCT images got noisier when the number of projections was reduced, but their spatial resolution and uniformity were hardly affected. The maximum registration errors under the small amount transformation of the reference CT images were found to be within 0.7 mm translation and 0.3 masculine rotation. However, when the projection number was lower than one-fourth of the full set with a large amount of transformation of reference CT images, the registration could easily be trapped into local minima solutions for a nonrigid anatomy. We concluded, by using projection-number reduction strategy under conscientious care, imaging-guided localization procedure could achieve a lower patient dose without losing the registration accuracy for various clinical sites and situations. A faster scanning time is the main advantage compared to the mA decrease-based, dose-reduction method.

  10. Influence of chamber misalignment on cased telescoped (CT ammunition accuracy

    Directory of Open Access Journals (Sweden)

    D. Corriveau

    2016-04-01

    Full Text Available As part of a research program, it was desired to better understand the impact of the rotating chamber alignment with the barrel throat on the precision and accuracy of a novel cased telescoped (CT ammunition firing rifle. In order to perform the study, a baseline CT ammunition chamber which was concentric with a Mann barrel bore was manufactured. Additionally, six chambers were manufactured with an offset relative to the barrel bore. These chambers were used to simulate a misaligned chamber relative to the bore axis. Precision and accuracy tests were then performed at 200 m in an indoor range under controlled conditions. For this project, 5.56 mm CT ammunition was used. As the chamber axis offset relative to the gun bore was increased, the mean point of impact was displaced away from the target center. The shift in the impact location is explained by the presence of in-bore yaw which results in lateral throw-off and aerodynamic jump components. The linear theory of ballistics is used to establish a relationship between the chamber misalignment and the resulting projectile mean point of impact for a rifle developed to fire CT ammunition. This relationship allows for the prediction of the mean point of impact given a chamber misalignment.

  11. Accuracy and Radiation Dose of CT-Based Attenuation Correction for Small Animal PET: A Monte Carlo Simulation Study

    International Nuclear Information System (INIS)

    Yang, Ching-Ching; Chan, Kai-Chieh

    2013-06-01

    -Small animal PET allows qualitative assessment and quantitative measurement of biochemical processes in vivo, but the accuracy and reproducibility of imaging results can be affected by several parameters. The first aim of this study was to investigate the performance of different CT-based attenuation correction strategies and assess the resulting impact on PET images. The absorbed dose in different tissues caused by scanning procedures was also discussed to minimize biologic damage generated by radiation exposure due to PET/CT scanning. A small animal PET/CT system was modeled based on Monte Carlo simulation to generate imaging results and dose distribution. Three energy mapping methods, including the bilinear scaling method, the dual-energy method and the hybrid method which combines the kVp conversion and the dual-energy method, were investigated comparatively through assessing the accuracy of estimating linear attenuation coefficient at 511 keV and the bias introduced into PET quantification results due to CT-based attenuation correction. Our results showed that the hybrid method outperformed the bilinear scaling method, while the dual-energy method achieved the highest accuracy among the three energy mapping methods. Overall, the accuracy of PET quantification results have similar trend as that for the estimation of linear attenuation coefficients, whereas the differences between the three methods are more obvious in the estimation of linear attenuation coefficients than in the PET quantification results. With regards to radiation exposure from CT, the absorbed dose ranged between 7.29-45.58 mGy for 50-kVp scan and between 6.61-39.28 mGy for 80-kVp scan. For 18 F radioactivity concentration of 1.86x10 5 Bq/ml, the PET absorbed dose was around 24 cGy for tumor with a target-to-background ratio of 8. The radiation levels for CT scans are not lethal to the animal, but concurrent use of PET in longitudinal study can increase the risk of biological effects. The

  12. Iterative reconstruction technique with reduced volume CT dose index: diagnostic accuracy in pediatric acute appendicitis

    International Nuclear Information System (INIS)

    Didier, Ryne A.; Vajtai, Petra L.; Hopkins, Katharine L.

    2015-01-01

    Iterative reconstruction technique has been proposed as a means of reducing patient radiation dose in pediatric CT. Yet, the effect of such reductions on diagnostic accuracy has not been thoroughly evaluated. This study compares accuracy of diagnosing pediatric acute appendicitis using contrast-enhanced abdominopelvic CT scans performed with traditional pediatric weight-based protocols and filtered back projection reconstruction vs. a filtered back projection/iterative reconstruction technique blend with reduced volume CT dose index (CTDI vol ). Results of pediatric contrast-enhanced abdominopelvic CT scans done for pain and/or suspected appendicitis were reviewed in two groups: A, 192 scans performed with the hospital's established weight-based CT protocols and filtered back projection reconstruction; B, 194 scans performed with iterative reconstruction technique and reduced CTDI vol . Reduced CTDI vol was achieved primarily by reductions in effective tube current-time product (mAs eff ) and tube peak kilovoltage (kVp). CT interpretation was correlated with clinical follow-up and/or surgical pathology. CTDI vol , size-specific dose estimates (SSDE) and performance characteristics of the two CT techniques were then compared. Between groups A and B, mean CTDI vol was reduced by 45%, and mean SSDE was reduced by 46%. Sensitivity, specificity and diagnostic accuracy were 96%, 97% and 96% in group A vs. 100%, 99% and 99% in group B. Accuracy in diagnosing pediatric acute appendicitis was maintained in contrast-enhanced abdominopelvic CT scans that incorporated iterative reconstruction technique, despite reductions in mean CTDI vol and SSDE by nearly half as compared to the hospital's traditional weight-based protocols. (orig.)

  13. Evaluation of solitary pulmonary nodules by integrated PET/CT: improved accuracy by FDG uptake pattern and CT findings

    International Nuclear Information System (INIS)

    Joon Young Choi; Kyung Soo Lee; O Jung Kwon; Young Mog Shim; Kyung-Han Lee; Yong Choi; Yearn Seong Choe; Byung-Tae Kim

    2004-01-01

    Objective: FDG PET is useful to differentiate malignancy from benign lesions in the evaluation of solitary pulmonary nodules (SPNs). However, FDG PET showed false positive results in benign inflammatory lesions such as tuberculosis and organizing pneumonia. Furthermore, malignant tumors such as adenocarcinoma (AC) with bronchioloalveolar carcinoma (BAC) type had lower FDG uptake than other cell types of non-small cell lung cancer. We investigated whether FDG uptake pattern and image findings of CT for attenuation correction could improve accuracy for evaluating SPNs over SUV in integrated PET/CT imaging using FDG. Methods: Forty patients (M:F = 23:17, mean age 58.2±9.4 yrs) with non-calcified SPNs (diameter on CT 30 mm, no significant mediastinal node enlargement, no atelectasis) were included. All subjects underwent integrated PET/CT imaging using FDG. One nuclear medicine physician and 1 chest radiologist interpreted the PET and non-contrast CT images for attenuation correction, respectively. On PET images, maximum SUV of SPN was acquired, and FDG uptake pattern was categorized as diffusely increased or heterogeneously increased with upper threshold of window setting adjusted to maximum SUV of each nodule. A radiologist interpreted SPNs as benign or malignant based on CT images with lung and mediastinai window settings blinded to PET findings. Results: On pathological exam, 30 SPNs were confirmed to be malignant (11 AC with non-BAC type, 8 AC with BAC type, 8 squamous cell carcinoma, 1 adenosquamous cell carcinoma, 1 neuroendocrine carcinoma, 1 large cell carcinoma), and 10 were benign (4 tuberculosis, 3 organizing pneumonia, 2 sclerosing pneumocytoma, 1 non-specific inflammation). All 5 nodules with max SUV 7.0 except one with tuberculoma had malignancy. When only nodules with diffusely increased uptake were considered malignant in indeterminate group with max SUV of 4.0 to 7.0, PET could diagnose 5 of 9 malignant nodules with one false positive nodule. In 6 of

  14. Accuracy of cancellous bone volume fraction measured by micro-CT scanning

    DEFF Research Database (Denmark)

    Ding, Ming; Odgaard, A; Hvid, I

    1999-01-01

    Volume fraction, the single most important parameter in describing trabecular microstructure, can easily be calculated from three-dimensional reconstructions of micro-CT images. This study sought to quantify the accuracy of this measurement. One hundred and sixty human cancellous bone specimens...... which covered a large range of volume fraction (9.8-39.8%) were produced. The specimens were micro-CT scanned, and the volume fraction based on Archimedes' principle was determined as a reference. After scanning, all micro-CT data were segmented using individual thresholds determined by the scanner...

  15. Reassessment of CT images to improve diagnostic accuracy in patients with suspected acute appendicitis and an equivocal preoperative CT interpretation

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Hyun Cheol; Yang, Dal Mo; Kim, Sang Won [Kyung Hee University Hospital at Gangdong, College of Medicine, Kyung Hee University, Department of Radiology, Seoul (Korea, Republic of); Park, Seong Jin [Kyung Hee University Hospital, College of Medicine, Kyung Hee University, Department of Radiology, Seoul (Korea, Republic of)

    2012-06-15

    To identify CT features that discriminate individuals with and without acute appendicitis in patients with equivocal CT findings, and to assess whether knowledge of these findings improves diagnostic accuracy. 53 patients that underwent appendectomy with an indeterminate preoperative CT interpretation were selected and allocated to an acute appendicitis group or a non-appendicitis group. The 53 CT examinations were reviewed by two radiologists in consensus to identify CT findings that could aid in the discrimination of those with and without appendicitis. In addition, two additional radiologists were then requested to evaluate independently the 53 CT examinations using a 4-point scale, both before and after being informed of the potentially discriminating criteria. CT findings found to be significantly different in the two groups were; the presence of appendiceal wall enhancement, intraluminal air in appendix, a coexistent inflammatory lesion, and appendiceal wall thickening (P < 0.05). Areas under the curves of reviewers 1 and 2 significantly increased from 0.516 and 0.706 to 0.677 and 0.841, respectively, when reviewers were told which CT variables were significant (P = 0.0193 and P = 0.0397, respectively). Knowledge of the identified CT findings was found to improve diagnostic accuracy for acute appendicitis in patients with equivocal CT findings. circle Numerous patients with clinically equivocal appendicitis do not have acute appendicitis circle Computed tomography (CT) helps to reduce the negative appendectomy rate circle CT is not always infallible and may also demonstrate indeterminate findings circle However knowledge of significant CT variables can further reduce negative appendectomy rate circle An equivocal CT interpretation of appendicitis should be reassessed with this knowledge. (orig.)

  16. Reassessment of CT images to improve diagnostic accuracy in patients with suspected acute appendicitis and an equivocal preoperative CT interpretation

    International Nuclear Information System (INIS)

    Kim, Hyun Cheol; Yang, Dal Mo; Kim, Sang Won; Park, Seong Jin

    2012-01-01

    To identify CT features that discriminate individuals with and without acute appendicitis in patients with equivocal CT findings, and to assess whether knowledge of these findings improves diagnostic accuracy. 53 patients that underwent appendectomy with an indeterminate preoperative CT interpretation were selected and allocated to an acute appendicitis group or a non-appendicitis group. The 53 CT examinations were reviewed by two radiologists in consensus to identify CT findings that could aid in the discrimination of those with and without appendicitis. In addition, two additional radiologists were then requested to evaluate independently the 53 CT examinations using a 4-point scale, both before and after being informed of the potentially discriminating criteria. CT findings found to be significantly different in the two groups were; the presence of appendiceal wall enhancement, intraluminal air in appendix, a coexistent inflammatory lesion, and appendiceal wall thickening (P < 0.05). Areas under the curves of reviewers 1 and 2 significantly increased from 0.516 and 0.706 to 0.677 and 0.841, respectively, when reviewers were told which CT variables were significant (P = 0.0193 and P = 0.0397, respectively). Knowledge of the identified CT findings was found to improve diagnostic accuracy for acute appendicitis in patients with equivocal CT findings. circle Numerous patients with clinically equivocal appendicitis do not have acute appendicitis circle Computed tomography (CT) helps to reduce the negative appendectomy rate circle CT is not always infallible and may also demonstrate indeterminate findings circle However knowledge of significant CT variables can further reduce negative appendectomy rate circle An equivocal CT interpretation of appendicitis should be reassessed with this knowledge. (orig.)

  17. Comparison of CT number calibration techniques for CBCT-based dose calculation

    International Nuclear Information System (INIS)

    Dunlop, Alex; McQuaid, Dualta; Nill, Simeon; Hansen, Vibeke N.; Oelfke, Uwe; Murray, Julia; Bhide, Shreerang; Harrington, Kevin; Poludniowski, Gavin; Nutting, Christopher; Newbold, Kate

    2015-01-01

    The aim of this work was to compare and validate various computed tomography (CT) number calibration techniques with respect to cone beam CT (CBCT) dose calculation accuracy. CBCT dose calculation accuracy was assessed for pelvic, lung, and head and neck (H and N) treatment sites for two approaches: (1) physics-based scatter correction methods (CBCT r ); (2) density override approaches including assigning water density to the entire CBCT (W), assignment of either water or bone density (WB), and assignment of either water or lung density (WL). Methods for CBCT density assignment within a commercially available treatment planning system (RS auto ), where CBCT voxels are binned into six density levels, were assessed and validated. Dose-difference maps and dose-volume statistics were used to compare the CBCT dose distributions with the ground truth of a planning CT acquired the same day as the CBCT. For pelvic cases, all CTN calibration methods resulted in average dose-volume deviations below 1.5 %. RS auto provided larger than average errors for pelvic treatments for patients with large amounts of adipose tissue. For H and N cases, all CTN calibration methods resulted in average dose-volume differences below 1.0 % with CBCT r (0.5 %) and RS auto (0.6 %) performing best. For lung cases, WL and RS auto methods generated dose distributions most similar to the ground truth. The RS auto density override approach is an attractive option for CTN adjustments for a variety of anatomical sites. RS auto methods were validated, resulting in dose calculations that were consistent with those calculated on diagnostic-quality CT images, for CBCT images acquired of the lung, for patients receiving pelvic RT in cases without excess adipose tissue, and for H and N cases. (orig.) [de

  18. Iterative reconstruction technique with reduced volume CT dose index: diagnostic accuracy in pediatric acute appendicitis

    Energy Technology Data Exchange (ETDEWEB)

    Didier, Ryne A. [Oregon Health and Science University, Department of Diagnostic Radiology, DC7R, Portland, OR (United States); Vajtai, Petra L. [Oregon Health and Science University, Department of Pediatrics, Portland, OR (United States); Oregon Health and Science University, Department of Diagnostic Radiology, DC7R, Portland, OR (United States); Hopkins, Katharine L. [Oregon Health and Science University, Department of Diagnostic Radiology, DC7R, Portland, OR (United States); Oregon Health and Science University, Department of Pediatrics, Portland, OR (United States)

    2014-07-05

    Iterative reconstruction technique has been proposed as a means of reducing patient radiation dose in pediatric CT. Yet, the effect of such reductions on diagnostic accuracy has not been thoroughly evaluated. This study compares accuracy of diagnosing pediatric acute appendicitis using contrast-enhanced abdominopelvic CT scans performed with traditional pediatric weight-based protocols and filtered back projection reconstruction vs. a filtered back projection/iterative reconstruction technique blend with reduced volume CT dose index (CTDI{sub vol}). Results of pediatric contrast-enhanced abdominopelvic CT scans done for pain and/or suspected appendicitis were reviewed in two groups: A, 192 scans performed with the hospital's established weight-based CT protocols and filtered back projection reconstruction; B, 194 scans performed with iterative reconstruction technique and reduced CTDI{sub vol}. Reduced CTDI{sub vol} was achieved primarily by reductions in effective tube current-time product (mAs{sub eff}) and tube peak kilovoltage (kVp). CT interpretation was correlated with clinical follow-up and/or surgical pathology. CTDI{sub vol}, size-specific dose estimates (SSDE) and performance characteristics of the two CT techniques were then compared. Between groups A and B, mean CTDI{sub vol} was reduced by 45%, and mean SSDE was reduced by 46%. Sensitivity, specificity and diagnostic accuracy were 96%, 97% and 96% in group A vs. 100%, 99% and 99% in group B. Accuracy in diagnosing pediatric acute appendicitis was maintained in contrast-enhanced abdominopelvic CT scans that incorporated iterative reconstruction technique, despite reductions in mean CTDI{sub vol} and SSDE by nearly half as compared to the hospital's traditional weight-based protocols. (orig.)

  19. Neuronavigation accuracy dependence on CT and MR imaging parameters: a phantom-based study

    International Nuclear Information System (INIS)

    Poggi, S; Pallotta, S; Russo, S; Gallina, P; Torresin, A; Bucciolini, M

    2003-01-01

    Clinical benefits from neuronavigation are well established. However, the complexity of its technical environment requires a careful evaluation of different types of errors. In this work, a detailed phantom study which investigates the accuracy in a neuronavigation procedure is presented. The dependence on many different imaging parameters, such as field of view, slice thickness and different kind of sequences (sequential and spiral for CT, T1-weighted and T2-weighted for MRI), is quantified. Moreover, data based on CT images are compared to those based on MR images, taking into account MRI distortion. Finally, the contributions to global accuracy coming from image acquisition, registration and navigation itself are discussed. Results demonstrate the importance of imaging accuracy. Procedures based on CT proved to be more accurate than procedures based on MRI. In the former, values from 2 to 2.5 mm are obtained for 95% fractiles of cumulative distribution of Euclidean distances between the intended target and the reached one while, in the latter, the measured values range from 3 to 4 mm. The absence of imaging distortion proved to be crucial for registration accuracy in MR-based procedures

  20. Accuracy improvement of CT reconstruction using tree-structured filter bank

    International Nuclear Information System (INIS)

    Ueda, Kazuhiro; Morimoto, Hiroaki; Morikawa, Yoshitaka; Murakami, Junichi

    2009-01-01

    Accuracy improvement of 'CT reconstruction algorithm using TSFB (Tree-Structured Filter Bank)' that is high-speed CT reconstruction algorithm, was proposed. TSFB method could largely reduce the amount of computation in comparison with the CB (Convolution Backprojection) method, but it was the problem that an artifact occurred in a reconstruction image since reconstruction was performed with disregard to a signal out of the reconstruction domain in stage processing. Also the whole band filter being the component of a two-dimensional synthesis filter was IIR filter and then an artifact occurred at the end of the reconstruction image. In order to suppress these artifacts the proposed method enlarged the processing range by the TSFB method in the domain outside by the width control of the specimen line and line addition to the reconstruction domain outside. And, furthermore, to avoid increase of the amount of computation, the algorithm was proposed such as to decide the needed processing range depending on the number of steps processing with the TSFB and the degree of incline of filter, and then update the position and width of the specimen line to process the needed range. According to the simulation to realize a high-speed and highly accurate CT reconstruction in this way, the quality of the reconstruction image of the proposed method was improved in comparison with the TSFB method and got the same result with the CB method. (T. Tanaka)

  1. Accuracy of 18F-FDG PET/CT for lymph node staging in non-small-cell lung cancers

    Institute of Scientific and Technical Information of China (English)

    LIU Bao-jun; DONG Jing-cheng; XU Chang-qing; ZUO Chuan-tao; LE Jing-jing; GUAN Yi-hui; ZHAO Jun; WU Jin-feng; DUAN Xiao-hong; CAO Yu-xue

    2009-01-01

    Background This retrospective study evaluated the diagnostic accuracy of 2-(F18)-fluoro-2-deoxy-D-glucose-positron emission tomography(18F-FDG-PET)/COmputed tomography(PET/CT)in the preoperative diagnosis of metastatic mediastinal and hilar lymph node in patients with non-small-cell lung cancer(NSCLC).Methods A total of 39 patients received preoperative 18F-FDG PET/CT and the postoperative biopsy.We compared preoperative PET/CT scan results with corresponding intraoperative histopathalogic findings in 39 NSCLC patients.The sensitivity,specificity,accuracy,positive and negative predictive value of 18F-FDG PET/CT were assessed.Results Histopathologic examination confirmed metastasis in 57 out of the 208 excised lymph nodes;23 of the 57 nodes were mediastinal and hilar lymph nodes.The sensitivity,specificity,accuracy,positive predictive value and negative predictive value of PET/CT in the preoperative diagnosis of mediastinal lymph node metastasis in NSCLC patients were 65%,96.8%,92%,78.5%and 90%,respectively.Conclusions PET/CT scan showed good accuracy in the preoperative diagnosis of mediastinal and hilar lymph node metastasis in the patients with NSCLC.We recommend that PET/CT scanning be used as a first-line evaluation tool for tumor diagnosis,therapy evaluation and follow-up.

  2. Characterization of the relation between CT technical parameters and accuracy of quantification of lung attenuation on quantitative chest CT.

    Science.gov (United States)

    Trotta, Brian M; Stolin, Alexander V; Williams, Mark B; Gay, Spencer B; Brody, Alan S; Altes, Talissa A

    2007-06-01

    The purpose of this study was to assess the compromise between CT technical parameters and the accuracy of CT quantification of lung attenuation. Materials that simulate water (0 H), healthy lung (-650 H), borderline emphysematous lung (-820 H), and severely emphysematous lung (-1,000 H) were placed at both the base and the apex of the lung of an anthropomorphic phantom and outside the phantom. Transaxial CT images through the samples were obtained while the effective tube current was varied from 440 to 10 mAs, kilovoltage from 140 to 80 kVp, and slice thickness from 0.625 to 10 mm. Mean +/- SD attenuation within the samples and the standard quantitative chest CT measurements, the percentage of pixels with attenuation less than -910 H and 15th percentile of attenuation, were computed. Outside the phantom, variations in CT parameters produced less than 2.0% error in all measurements. Within the anthropomorphic phantom at 30 mAs, error in measurements was much larger, ranging from zero to 200%. Below approximately 80 mAs, mean attenuation became increasingly biased. The effects were most pronounced at the apex of the lungs. Mean attenuation of the borderline emphysematous sample of apex decreased 55 H as the tube current was decreased from 300 to 30 mAs. Both the 15th percentile of attenuation and percentage of pixels with less than -910 H attenuation were more sensitive to variations in effective tube current than was mean attenuation. For example, the -820 H sample should have 0% of pixels less than -910 H, which was true at 400 mA. At 30 mA in the lung apex, however, the measurement was highly inaccurate, 51% of pixels being below this value. Decreased kilovoltage and slice thickness had analogous, but lesser, effects. The accuracy of quantitative chest CT is determined by the CT acquisition parameters. There can be significant decreases in accuracy at less than 80 mAs for thin slices in an anthropomorphic phantom, the most pronounced effects occurring in the lung

  3. The accuracy of {sup 68}Ga-PSMA PET/CT in primary lymph node staging in high-risk prostate cancer

    Energy Technology Data Exchange (ETDEWEB)

    Oebek, Can; Doganca, Tuenkut [Acibadem Taksim Hospital, Department of Urology, Istanbul (Turkey); Demirci, Emre [Sisli Etfal Training and Research Hospital, Department of Nuclear Medicine, Istanbul (Turkey); Ocak, Meltem [Istanbul University, Faculty of Pharmacy, Department of Pharmaceutical Technology, Istanbul (Turkey); Kural, Ali Riza [Acibadem University, Department of Urology, Istanbul (Turkey); Yildirim, Asif [Istanbul Medeniyet University, Department of Urology, Istanbul (Turkey); Yuecetas, Ugur [Istanbul Training and Research Hospital, Department of Urology, Istanbul (Turkey); Demirdag, Cetin [Istanbul University, Cerrahpasa School of Medicine, Department of Urology, Istanbul (Turkey); Erdogan, Sarper M. [Istanbul University, Cerrahpasa School of Medicine, Department of Public Health, Istanbul (Turkey); Kabasakal, Levent [Istanbul University, Cerrahpasa School of Medicine, Department of Nuclear Medicine, Istanbul (Turkey); Collaboration: Members of Urooncology Association, Turkey

    2017-10-15

    To assess the diagnostic accuracy of {sup 68}Ga-PSMA PET in predicting lymph node (LN) metastases in primary N staging in high-risk and very high-risk nonmetastatic prostate cancer in comparison with morphological imaging. This was a multicentre trial of the Society of Urologic Oncology in Turkey in conjunction with the Nuclear Medicine Department of Cerrahpasa School of Medicine, Istanbul University. Patients were accrued from eight centres. Patients with high-risk and very high-risk disease scheduled to undergo surgical treatment with extended LN dissection between July 2014 and October 2015 were included. Either MRI or CT was used for morphological imaging. PSMA PET/CT was performed and evaluated at a single centre. Sensitivity, specificity and accuracy were calculated for the detection of lymphatic metastases by PSMA PET/CT and morphological imaging. Kappa values were calculated to evaluate the correlation between the numbers of LN metastases detected by PSMA PET/CT and by histopathology. Data on 51 eligible patients are presented. The sensitivity, specificity and accuracy of PSMA PET in detecting LN metastases in the primary setting were 53%, 86% and 76%, and increased to 67%, 88% and 81% in the subgroup with of patients with ≥15 LN removed. Kappa values for the correlation between imaging and pathology were 0.41 for PSMA PET and 0.18 for morphological imaging. PSMA PET/CT is superior to morphological imaging for the detection of metastatic LNs in patients with primary prostate cancer. Surgical dissection remains the gold standard for precise lymphatic staging. (orig.)

  4. CT-guided percutaneous transthoracic aspiration biopsy of lung lesions: factors influencing the diagnostic accuracy

    International Nuclear Information System (INIS)

    Huang Zhenguo; Zhang Xuezhe; Wang Wu

    2006-01-01

    Objective: To investigate the factors influencing the diagnostic accuracy in CT-guided percutaneous transthoracic aspiration biopsy of lung lesions. Methods: CT-guided percutaneous transthoracic biopsy was performed in 435 patients with surgical pathologic confirmation or clinical follow-up. Univariant statistical analysis and multivariant stepwise Logistic regression analysis were used to study the influence of lesion-related factors (benignancy or malignancy, size, location, depth, presence of cavity), procedure factors (laser-guidance, position of patient) and patient-related factors (sex, age, presence of emphysema); with the accuracy of the diagnosis. Results: 1)289 lesions were diagnosed as malignant ones and 146 as benign ones. The accuracy of CT-guided biopsy was 83.4%(363/435). Univariant analysis showed that the diagnostic accuracies for malignant and benign lesions were 88.9%(257/289) and 72.6%(106/146), respectively (χ 2 =18.71, P=0.00002). Laser-group was superior to non-laser group (88.4% versus 80.9%, respectively, χ 2 =4.00, P=0.0456). Mean diameters of lesions in correct diagnostic group and non-correct diagnostic group were 3.78±1.64 cm and 3.02±1.26 cm, respectively (F=13.79, P=0.0002). 2)Multivariant stepwise Logistic regression analysis showed that among the various factors influencing the final benign or malignant diagnoses (Wald χ 2 =14.01, P=0.0002) and using laser-guidance(Wald χ 2 =3.92, P=0.0477) were significantly associated with the diagnostic accuracy. Conclusions: Final correct diagnoses (benign, malignant) are closely related to the application of laser-guidance for determining the diagnostic accuracy in CT-guided transcutaneous thoracic needle aspiration biopsy. (authors)

  5. CT reconstruction techniques for improved accuracy of lung CT airway measurement

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez, A. [Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin 53705 (United States); Ranallo, F. N. [Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin 53705 and Department of Radiology, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin 53792 (United States); Judy, P. F. [Brigham and Women’s Hospital, Boston, Massachusetts 02115 (United States); Gierada, D. S. [Department of Radiology, Washington University, St. Louis, Missouri 63110 (United States); Fain, S. B., E-mail: sfain@wisc.edu [Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin 53705 (United States); Department of Radiology, University of Wisconsin School of Medicine and Public Health, Madison, Wisconsin 53792 (United States); Department of Biomedical Engineering,University of Wisconsin School of Engineering, Madison, Wisconsin 53706 (United States)

    2014-11-01

    FBP. Veo reconstructions showed slight improvement over STD FBP reconstructions (4%–9% increase in accuracy). The most improved ID and WA% measures were for the smaller airways, especially for low dose scans reconstructed at half DFOV (18 cm) with the EDGE algorithm in combination with 100% ASIR to mitigate noise. Using the BONE + ASIR at half BONE technique, measures improved by a factor of 2 over STD FBP even at a quarter of the x-ray dose. Conclusions: The flexibility of ASIR in combination with higher frequency algorithms, such as BONE, provided the greatest accuracy for conventional and low x-ray dose relative to FBP. Veo provided more modest improvement in qCT measures, likely due to its compatibility only with the smoother STD kernel.

  6. CT reconstruction techniques for improved accuracy of lung CT airway measurement

    International Nuclear Information System (INIS)

    Rodriguez, A.; Ranallo, F. N.; Judy, P. F.; Gierada, D. S.; Fain, S. B.

    2014-01-01

    FBP. Veo reconstructions showed slight improvement over STD FBP reconstructions (4%–9% increase in accuracy). The most improved ID and WA% measures were for the smaller airways, especially for low dose scans reconstructed at half DFOV (18 cm) with the EDGE algorithm in combination with 100% ASIR to mitigate noise. Using the BONE + ASIR at half BONE technique, measures improved by a factor of 2 over STD FBP even at a quarter of the x-ray dose. Conclusions: The flexibility of ASIR in combination with higher frequency algorithms, such as BONE, provided the greatest accuracy for conventional and low x-ray dose relative to FBP. Veo provided more modest improvement in qCT measures, likely due to its compatibility only with the smoother STD kernel

  7. Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

    International Nuclear Information System (INIS)

    Michalak, Gregory; Grimes, Joshua; Fletcher, Joel; Yu, Lifeng; Leng, Shuai; McCollough, Cynthia; Halaweish, Ahmed

    2016-01-01

    Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside ±10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors’ report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE

  8. Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

    Energy Technology Data Exchange (ETDEWEB)

    Michalak, Gregory; Grimes, Joshua; Fletcher, Joel; Yu, Lifeng; Leng, Shuai; McCollough, Cynthia, E-mail: mccollough.cynthia@mayo.edu [Department of Radiology, Mayo Clinic, Rochester, Minnesota 55905 (United States); Halaweish, Ahmed [Siemens Medical Solutions, Malvern, Pennsylvania 19355 (United States)

    2016-01-15

    Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside ±10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors’ report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE.

  9. Anastomotic leakage after colorectal surgery: diagnostic accuracy of CT

    Energy Technology Data Exchange (ETDEWEB)

    Kauv, Paul; Benadjaoud, Samir; Boulay-Coletta, Isabelle; Zins, Marc [Fondation Hopital Saint-Joseph, Department of Radiology, Paris (France); Curis, Emmanuel [Universite Paris Descartes, Laboratoire de biomathematiques, Faculte de pharmacie, Paris (France); Loriau, Jerome [Fondation Hopital Saint-Joseph, Department of Digestive Surgery, Paris (France)

    2015-12-15

    To evaluate the diagnostic accuracy of CT in postoperative colorectal anastomotic leakage (AL). Two independent blinded radiologists reviewed 153 CTs performed for suspected AL within 60 days after surgery in 131 consecutive patients, with (n = 58) or without (n = 95) retrograde contrast enema (RCE). Results were compared to original interpretations. The reference standard was reoperation or consensus (a radiologist and a surgeon) regarding clinical, laboratory, radiological, and follow-up data after medical treatment. AL was confirmed in 34/131 patients. For the two reviewers and original interpretation, sensitivity of CT was 82 %, 87 %, and 71 %, respectively; specificity was 84 %, 84 %, and 92 %. RCE significantly increased the positive predictive value (from 40 % to 88 %, P = 0.0009; 41 % to 92 %, P = 0.0016; and 40 % to 100 %, P = 0.0006). Contrast extravasation was the most sensitive (reviewers, 83 % and 83 %) and specific (97 % and 97 %) sign and was significantly associated with AL by univariate analysis (P < 0.0001 and P < 0.0001). By multivariate analysis with recursive partitioning, CT with RCE was accurate to confirm or rule out AL with contrast extravasation. CT with RCE is accurate for diagnosing postoperative colorectal AL. Contrast extravasation is the most reliable sign. RCE should be performed during CT for suspected AL. (orig.)

  10. Diagnostic accuracy of high-definition CT coronary angiography in high-risk patients

    International Nuclear Information System (INIS)

    Iyengar, S.S.; Morgan-Hughes, G.; Ukoumunne, O.; Clayton, B.; Davies, E.J.; Nikolaou, V.; Hyde, C.J.; Shore, A.C.; Roobottom, C.A.

    2016-01-01

    Aim: To assess the diagnostic accuracy of computed tomography coronary angiography (CTCA) using a combination of high-definition CT (HD-CTCA) and high level of reader experience, with invasive coronary angiography (ICA) as the reference standard, in high-risk patients for the investigation of coronary artery disease (CAD). Materials and methods: Three hundred high-risk patients underwent HD-CTCA and ICA. Independent experts evaluated the images for the presence of significant CAD, defined primarily as the presence of moderate (≥50%) stenosis and secondarily as the presence of severe (≥70%) stenosis in at least one coronary segment, in a blinded fashion. HD-CTCA was compared to ICA as the reference standard. Results: No patients were excluded. Two hundred and six patients (69%) had moderate and 178 (59%) had severe stenosis in at least one vessel at ICA. The sensitivity, specificity, positive predictive value, and negative predictive value were 97.1%, 97.9%, 99% and 93.9% for moderate stenosis, and 98.9%, 93.4%, 95.7% and 98.3%, for severe stenosis, on a per-patient basis. Conclusion: The combination of HD-CTCA and experienced readers applied to a high-risk population, results in high diagnostic accuracy comparable to ICA. Modern generation CT systems in experienced hands might be considered for an expanded role. - Highlights: • Diagnostic accuracy of High-Definition CT Angiography (HD-CTCA) has been assessed. • Invasive Coronary angiography (ICA) is the reference standard. • Diagnostic accuracy of HD-CTCA is comparable to ICA. • Diagnostic accuracy is not affected by coronary calcium or stents. • HD-CTCA provides a non-invasive alternative in high-risk patients.

  11. Estimation of regional intrapulmonary moisture volumes by pulmonary CT numbers in CT views

    International Nuclear Information System (INIS)

    Nakamoto, Takaaki; Kato, Shiro; Yoshimura, Masaharu

    1987-01-01

    In order to study the regional intrapulmonary moisture volume in normal lungs, lungs with congestion, and lungs with increased blood flow, a series of 42 subjects underwent transverse computed tomography (CT) scanning performed at a total of 17 points corresponding to the upper (7), middle (3), and lower (7) lung fields. The subjects were divided into group I (n = 24, with left-sided heart failure), group II (n = 8, with left-to-right shunt heart disease), and group III (n = 10, normals). In group I, the pulmonary peripheral minus CT numbers were inversely correlated with the pulmonary wedge mean pressure and pulmonary arterial mean pressure; and were linearly correlated with pulmonary blood flow (PBF). In group II, they were inversely correlated with PBF and left-to-right shunt volume. The pulmonary pheripheral CT numbers were significantly higher in both groups I and II than group III. In group III, these CT numbers were higher in the posterior field than the anterior field in supine position. (Hou, H.L.)

  12. Accuracy of CT enterography and magnetic resonance enterography imaging to detect lesions preoperatively in patients undergoing surgery for Crohn's disease.

    Science.gov (United States)

    Seastedt, Kenneth P; Trencheva, Koiana; Michelassi, Fabrizio; Alsaleh, Doaa; Milsom, Jeffrey W; Sonoda, Toyooki; Lee, Sang W; Nandakumar, Govind

    2014-12-01

    CT enterography and magnetic resonance enterography have emerged as first-line imaging technologies for the evaluation of the gastrointestinal tract in Crohn's disease. The purpose of this work was to evaluate the accuracy of these imaging modalities to identify Crohn's disease lesions preoperatively. This was a retrospective chart review. The study was conducted at a single institution. Seventy-six patients with Crohn's disease with preoperative CT enterography and/or magnetic resonance enterography were included in the study. The number of stenoses, fistulas, and abscesses on CT enterography and/or magnetic resonance enterography before surgery were compared with operative findings. Forty patients (53%) were women, 46 (60%) underwent surgery for recurrent Crohn's disease, and 46 (57%) had previous abdominal surgery. Thirty-six (47%) had a preoperative CT enterography and 43 (57%) had a preoperative magnetic resonance enterography. CT enterography sensitivity was 75% for stenosis and 50% for fistula. MRE sensitivity was 68% for stenosis and 60% for fistula. The negative predictive values of CT enterography and magnetic resonance enterography for stenosis were very low (54% and 65%) and were 85% and 81% for fistula. CT enterography had 76% accuracy for stenosis and 79% for fistula; magnetic resonance enterography had 78% accuracy for stenosis and 85% for fistula. Both were accurate for abscess. False-negative rates for CT enterography were 50% for fistula and 25% for stenosis. False-negative rates for magnetic resonance enterography were 40% for fistula and 32% for stenosis. Unexpected intraoperative findings led to modification of the planned surgical procedure in 20 patients (26%). This study was limited by its small sample size, its retrospective nature, and that some studies were performed at outside institutions. CT enterography and magnetic resonance enterography in patients with Crohn's disease were accurate for the identification of abscesses but not for

  13. CT fluoroscopy-guided vs. multislice CT biopsy mode-guided lung biopsies: Accuracy, complications and radiation dose

    International Nuclear Information System (INIS)

    Prosch, Helmut; Stadler, Alfred; Schilling, Matthias; Bürklin, Sandra; Eisenhuber, Edith; Schober, Ewald; Mostbeck, Gerhard

    2012-01-01

    Background: The aim of this retrospective study was to compare the diagnostic accuracy, the frequency of complications, the duration of the interventions and the radiation doses of CT fluoroscopy (CTF) guided biopsies of lung lesions with those of multislice CT (MS-CT) biopsy mode-guided biopsies. Methods: Data and images from 124 consecutive patients undergoing CTF-guided lung biopsy (group A) and 132 MS-CT-biopsy mode-guided lung biopsy (group B) were reviewed. CTF-guided biopsies were performed on a Siemens Emotion 6 CT scanner with intermittent or continuous CT-fluoroscopy, MS-CT biopsy mode-guided biopsies were performed on a Siemens Emotion 16 CT scanner. All biopsies were performed with a coaxial needle technique. Results: The two groups (A vs. B) did not differ significantly regarding sensitivity (95.5% vs. 95.9%), specificity (96.7% vs. 95.5%), negative predictive value (87.9% vs. 84%) or positive predictive value (98.8% vs. 98.9%). Pneumothorax was observed in 30.0% and 32.5% of the patients, respectively. Chest tube placement was necessary in 4% (group A) and 13% (group B) of the patients. The duration of the intervention was significantly longer in group A (median 37 min vs. 32 min, p = 0.04). The mean CT dose index (CTDI) was 422 in group A and 36.3 in group B (p < 0.001). Conclusion: Compared to CTF-guided biopsies, chest biopsies using the MS-CT biopsy mode show dramatically lower CTDI levels. Although the diagnostic yield of the procedures do not differ significantly, biopsies using the MS-CT-biopsy mode have a three-fold higher rate of chest tube placement.

  14. Clinical studies on thyroid CT number in Graves' disease and destructive thyrotoxicosis

    International Nuclear Information System (INIS)

    Kamijo, Keiichi

    1994-01-01

    The purpose of the present study was to investigate CT Hounsfield unit (H.U.) of the thyroid in hyperthyroid and euthyroid Graves' disease and destructive thyrotoxicosis. The mean thyroid CT number in 95 controls was 122±18 H.U.(±SD) and did not change significatly with advancing age. The mean thyroid CT number (±SD) of 85±22 H.U. in 60 patients with hyperthyroid Graves' disease was significantly lower than either in normal controls or 116±22 H.U. in 11 patients with euthyroid Graves' disease. Comparison of thyroid hormones and TSH receptor Ab values of untreated patients with a normal and an abnormally low thyroid CT number showed that serum total and free T 3 were significantly higher in the latter group than the former group. With respect to the effect of methimazol (MMI) on the thyroid CT number, in the untreated 10 patients with a low thyroid CT number, the initial mean CT number was 65±11 H.U. and increased significantly to 76±14 H.U. after treatment with MMI. In contrast, in 6 patients with a normal thyroid CT number prior to therapy, the initial mean thyroid CT number was 102±11 H.U. and fell significantly to 84±16 H.U. after treatment with MMI. The thyroid CT number in destructive thyrotoxicosis is markedly decreased to less than 70 H.U. and the mean values of 57±7 H.U. in 6 patients with silent thyroiditis and of 61±5 H.U. in 7 with subacute thyroiditis differ significantly from Graves' disease. In conclusion, the thyroid CT number is significantly reduced in hyperthyroid Graves' disease, normal in euthyroid Graves' disease and markedly decreased in destructive thyrotoxicosis. The high T-3 value seemed to play an important role in the pathogenesis of a decline in the thyroid CT number in Graves' disease. An antithyroid drug therapy caused two different changes in the thyroid CT number, depending on whether the thyroid CT number prior to therapy was normal or low. (author)

  15. Monte Carlo calculated CT numbers for improved heavy ion treatment planning

    Directory of Open Access Journals (Sweden)

    Qamhiyeh Sima

    2014-03-01

    Full Text Available Better knowledge of CT number values and their uncertainties can be applied to improve heavy ion treatment planning. We developed a novel method to calculate CT numbers for a computed tomography (CT scanner using the Monte Carlo (MC code, BEAMnrc/EGSnrc. To generate the initial beam shape and spectra we conducted full simulations of an X-ray tube, filters and beam shapers for a Siemens Emotion CT. The simulation output files were analyzed to calculate projections of a phantom with inserts. A simple reconstruction algorithm (FBP using a Ram-Lak filter was applied to calculate the pixel values, which represent an attenuation coefficient, normalized in such a way to give zero for water (Hounsfield unit (HU. Measured and Monte Carlo calculated CT numbers were compared. The average deviation between measured and simulated CT numbers was 4 ± 4 HU and the standard deviation σ was 49 ± 4 HU. The simulation also correctly predicted the behaviour of H-materials compared to a Gammex tissue substitutes. We believe the developed approach represents a useful new tool for evaluating the effect of CT scanner and phantom parameters on CT number values.

  16. Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris

    Energy Technology Data Exchange (ETDEWEB)

    Pugliese, Francesca; Krestin, Gabriel P. [Erasmus Medical Center, Department of Radiology, Rotterdam (Netherlands); Mollet, Nico R.A.; deFeyter, Pim J. [Erasmus Medical Center, Department of Radiology, Rotterdam (Netherlands); Erasmus Medical Center, Department of Cardiology, Rotterdam (Netherlands); Runza, Giuseppe [University of Palermo, Department of Radiology, Palermo (Italy); Azienda Ospedaliera di Parma, Department of Radiology, Parma (Italy); Mieghem, Carlos van; Meijboom, Willem B.; Baks, Timo [Erasmus Medical Center, Department of Cardiology, Rotterdam (Netherlands); Malagutti, Patrizia [Erasmus Medical Center, Department of Cardiology, Rotterdam (Netherlands); University of Ferrara, Department of Cardiology, Ferrara (Italy); Cademartiri, Filippo [Erasmus Medical Center, Department of Radiology, Rotterdam (Netherlands); Azienda Ospedaliera di Parma, Department of Radiology, Parma (Italy)

    2006-03-15

    Multislice computed tomography (CT) is an emerging technique for the non-invasive detection of coronary stenoses. While the diagnostic accuracy of 4-slice scanners was limited, 16-slice CT imagers showed promising results due to increased temporal and spatial resolution. These technical advances prompted us to evaluate the diagnostic performance of 64-slice CT coronary angiography in the detection of significant stenoses (defined as {>=} 50% luminal diameter reduction) versus invasive quantitative coronary angiography (QCA). Thirty-five patients with stable angina pectoris underwent CT coronary angiography performed with a 64-slice scanner (gantry rotation time 330 ms, individual detector width 0.6 mm) prior to conventional coronary angiography. Patients with heart rates >70 beats/min received 100 mg metoprolol orally. One hundred millilitres of contrast agent with an iodine concentration of 400 mgl/ml were injected at a rate of 5 ml/s into the antecubital vein. The CT scan was triggered with the bolus tracking technique. The sensitivity, specificity and the positive and negative predictive values of 64-slice CT were 99%, 96%, 78% and 99%, respectively, on a per-segment basis. The values obtained on a per-patient basis were 100%, 90%, 96% and 100%, respectively. When referral to catheterisation is questionable, CT coronary angiography may identify subjects with normal angiograms and consistently decrease the number of unnecessary invasive procedures. (orig.)

  17. Diagnostic accuracy of non-invasive 64-slice CT coronary angiography in patients with stable angina pectoris

    International Nuclear Information System (INIS)

    Pugliese, Francesca; Krestin, Gabriel P.; Mollet, Nico R.A.; DeFeyter, Pim J.; Runza, Giuseppe; Mieghem, Carlos van; Meijboom, Willem B.; Baks, Timo; Malagutti, Patrizia; Cademartiri, Filippo

    2006-01-01

    Multislice computed tomography (CT) is an emerging technique for the non-invasive detection of coronary stenoses. While the diagnostic accuracy of 4-slice scanners was limited, 16-slice CT imagers showed promising results due to increased temporal and spatial resolution. These technical advances prompted us to evaluate the diagnostic performance of 64-slice CT coronary angiography in the detection of significant stenoses (defined as ≥ 50% luminal diameter reduction) versus invasive quantitative coronary angiography (QCA). Thirty-five patients with stable angina pectoris underwent CT coronary angiography performed with a 64-slice scanner (gantry rotation time 330 ms, individual detector width 0.6 mm) prior to conventional coronary angiography. Patients with heart rates >70 beats/min received 100 mg metoprolol orally. One hundred millilitres of contrast agent with an iodine concentration of 400 mgl/ml were injected at a rate of 5 ml/s into the antecubital vein. The CT scan was triggered with the bolus tracking technique. The sensitivity, specificity and the positive and negative predictive values of 64-slice CT were 99%, 96%, 78% and 99%, respectively, on a per-segment basis. The values obtained on a per-patient basis were 100%, 90%, 96% and 100%, respectively. When referral to catheterisation is questionable, CT coronary angiography may identify subjects with normal angiograms and consistently decrease the number of unnecessary invasive procedures. (orig.)

  18. Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT

    NARCIS (Netherlands)

    Pelgrim, Gert Jan; van Hamersvelt, Robbert W; Willemink, Martin J; Schmidt, Bernhard T; Flohr, Thomas; Schilham, Arnold; Milles, Julien; Oudkerk, Matthijs; Leiner, Tim; Vliegenthart, Rozemarijn

    OBJECTIVE: To determine the accuracy of iodine quantification with dual energy computed tomography (DECT) in two high-end CT systems with different spectral imaging techniques. METHODS: Five tubes with different iodine concentrations (0, 5, 10, 15, 20 mg/ml) were analysed in an anthropomorphic

  19. Diagnostic accuracy and safety of CT-guided fine needle aspiration biopsy in cavitary pulmonary lesions

    Energy Technology Data Exchange (ETDEWEB)

    Zhuang, Yi-Ping, E-mail: yipingzhuang2010@sina.com [Department of Radiology, Jiangsu Cancer Institute and Hospital, No. 42 Baiziting Road, Nanjing 210009, Jiangsu (China); Wang, Hai-Yan, E-mail: mycherishgirl@sohu.com [Department of Radiology, Jiangsu Cancer Institute and Hospital, No. 42 Baiziting Road, Nanjing 210009, Jiangsu (China); Zhang, Jin, E-mail: yari_zj@hotmail.com [Department of Radiology, Jiangsu Cancer Institute and Hospital, No. 42 Baiziting Road, Nanjing 210009, Jiangsu (China); Feng, Yong, E-mail: fengyong119@sohu.com [Department of Radiology, Jiangsu Cancer Institute and Hospital, No. 42 Baiziting Road, Nanjing 210009, Jiangsu (China); Zhang, Lei, E-mail: motozl163@163.com [Department of Radiology, Jiangsu Cancer Institute and Hospital, No. 42 Baiziting Road, Nanjing 210009, Jiangsu (China)

    2013-01-15

    Objective: CT-guided transthoracic biopsy is a well-established method in the cytologic or histologic diagnosis of pulmonary lesions. The knowledge of its diagnostic performance and complications for cavitary pulmonary lesions is limited. The purpose of this study was to determine the diagnostic accuracy and safety of CT-guided fine needle aspiration biopsy (FNAB) in cavitary pulmonary lesions. Materials and methods: 102 consecutive patients with pulmonary cavitary lesions received CT-guided FNAB with use of an 18-gauge (n = 35) or 20-gauge (n = 67) Chiba for histology diagnosis. The sensitivity, specificity, and diagnostic accuracy of FNAB were calculated as compared with the final diagnosis. Complications associated with FNAB were observed. The diagnostic accuracy and complications were compared between patients with different lesion sizes and different cavity wall thickness. Results: The overall sensitivity, specificity, and accuracy of FNAB were 96.3%, 98.0%, and 96.1%, respectively. The sensitivity, specificity, and diagnosis accuracy in different lesion size (<2 cm vs ≥2 cm), or different cavity wall thickness (<5 mm vs ≥5 mm) were not different (P > 0.05; 0.235). More nondiagnostic sample was found in wall thickness <5 mm lesions (P = 0.017). Associated complications included pneumothorax in 9 (8.8%) patients and alveolar hemorrhage in 14 patients (13.7%) and hemoptysis in 1 patient (1%). No different rate of complications was found with regard to lesion size, wall thickness, length of the needle path and needle size (P > 0.05). Conclusion: CT-guided FNAB can be effectively ad safely used for patients with pulmonary cavitary lesions.

  20. Diagnostic accuracy of CT angiography in acute gastrointestinal bleeding

    International Nuclear Information System (INIS)

    Chua, A. E.; Ridley, L. J.

    2008-01-01

    Full text: The aim of the study was to carry out a systematic review determining the accuracy of CT angiography in the diagnosis of acute gastrointestinal bleeding. A search of published work in Medline and manual searching of reference lists of articles was conducted. Studies were included if they compared CT angiography to a reference standard of upper gastrointestinal endoscopy, colonoscopy, angiography or surgery in the diagnosis of acute gastrointestinal bleeding. Eight published studies evaluating 129 patients were included. Data were used to form 2 x2 tables. Computed tomography angiography showed pooled sensitivity of 86% (95% confidence interval 78-92%) and specificity of 95% (95% confidence interval 76-100%), without showing significant heterogeneity (x 2 = 3.5, P=0.6) and (x 2 - 5.4, P = 0.6), respectively. Summary receiver operating characteristic analysis showed an area under the curve of 0.93. Computed tomography angiography is accurate in the diagnosis of acute gastrointestinal bleeding and can show the precise location and aetiology of bleeding, thereby directing further management. Strong recommendations for use of CT cannot be made from this review because of the methodological limitations and further large prospective studies are needed to define the role of CT in acute gastrointestinal bleeding.

  1. Clinical studies on thyroid CT number in Graves' disease and destructive thyrotoxicosis

    Energy Technology Data Exchange (ETDEWEB)

    Kamijo, Keiichi (Kamijo Thyroid and Pituitary Clinic, Sapporo (Japan))

    1994-02-01

    The purpose of the present study was to investigate CT Hounsfield unit (H.U.) of the thyroid in hyperthyroid and euthyroid Graves' disease and destructive thyrotoxicosis. The mean thyroid CT number in 95 controls was 122[+-]18 H.U.([+-]SD) and did not change significatly with advancing age. The mean thyroid CT number ([+-]SD) of 85[+-]22 H.U. in 60 patients with hyperthyroid Graves' disease was significantly lower than either in normal controls or 116[+-]22 H.U. in 11 patients with euthyroid Graves' disease. Comparison of thyroid hormones and TSH receptor Ab values of untreated patients with a normal and an abnormally low thyroid CT number showed that serum total and free T[sub 3] were significantly higher in the latter group than the former group. With respect to the effect of methimazol (MMI) on the thyroid CT number, in the untreated 10 patients with a low thyroid CT number, the initial mean CT number was 65[+-]11 H.U. and increased significantly to 76[+-]14 H.U. after treatment with MMI. In contrast, in 6 patients with a normal thyroid CT number prior to therapy, the initial mean thyroid CT number was 102[+-]11 H.U. and fell significantly to 84[+-]16 H.U. after treatment with MMI. The thyroid CT number in destructive thyrotoxicosis is markedly decreased to less than 70 H.U. and the mean values of 57[+-]7 H.U. in 6 patients with silent thyroiditis and of 61[+-]5 H.U. in 7 with subacute thyroiditis differ significantly from Graves' disease. In conclusion, the thyroid CT number is significantly reduced in hyperthyroid Graves' disease, normal in euthyroid Graves' disease and markedly decreased in destructive thyrotoxicosis. The high T-3 value seemed to play an important role in the pathogenesis of a decline in the thyroid CT number in Graves' disease. An antithyroid drug therapy caused two different changes in the thyroid CT number, depending on whether the thyroid CT number prior to therapy was normal or low. (author).

  2. Diagnostic Accuracy of Detecting Hashimoto's Thyroiditis in Thyroid Cancer Patients Who Underwent Thyroid Surgery: Comparison of Ultrasonography, Positron Emission Tomography/CT, Contrast Enhanced CT, and Anti-Thyroid Antibody

    International Nuclear Information System (INIS)

    Kim, Young Gyun; Lee, Tae Hyun; Park, Dong Hee; Nam, Sang Been

    2012-01-01

    To compare the diagnostic accuracy of ultrasonography (US), F18-fluorodeoxyglucose positron emission tomography/CT (PET/CT), contrast enhanced CT (CECT), serum anti-thyroid antibody for detecting Hashimoto's thyroiditis in thyroid cancer patients who underwent neck surgery. A total of 150 patients with suspicious for thyroid cancer, who had previously undergone US guided needle aspiration of thyroid, were evaluated with the use of US, PET/CT, CECT and serum anti-thyroid antibody. The four studies were performed within two months before neck surgery. Hashimoto's thyroiditis was confirmed by histopathological results. The diagnostic accuracy of US, PET/CT, CECT and serum anti-thyroid antibody were calculated statistically. Hashimoto's thyroiditis was diagnosed in 51 out of the 150 patients, following neck surgery. The sensitivity, specificity, positive predictive value, negative predictive value and accuracy of US were 76.5%, 92.9%, 84.8%, 88.5%, and 87.3%, respectively. The corresponding values of PET/CT were 37.3%, 96.0%, 82.6%, 74.8%, and 76.0%, and CECT were 62.7%, 89.9%, 76.2%, 82.4%, and 80.7%, and serum anti-thyroid antibody level were 90.2%, 93.9%, 88.5%, 94.9%, and 92.7%, respectively. McNemar test revealed significant difference among PET/CT and others, but no significant differences among US, CECT and serum anti-thyroid antibody. Overall, serum anti-thyroid antibody showed most accurate diagnostic performance. In detecting Hashimoto's thyroiditis, serum anti-thyroid antibody showed higher diagnostic accuracy than others. US also showed relatively high diagnostic accuracy.

  3. Accuracy of FDG-PET-CT in the Diagnostic Work-up of Vascular Prosthetic Graft Infection

    NARCIS (Netherlands)

    Bruggink, J. L. M.; Glaudemans, A. W. J. M.; Saleem, B. R.; Meerwaldt, R.; Alkefaji, H.; Prins, T. R.; Start, R. H. J. A.; Zeebregts, C. J.

    Objectives: To investigate the diagnostic accuracy of fluoro-2-deoxy-D-glucose positron emission tomography (FDG-PET) compared with computed tomography (CT) scanning and added value of fused FDG-PET CT in diagnosing vascular prosthetic graft infection. Design: Prospective cohort study with

  4. Diagnostic Accuracy of MRI Versus CT for the Evaluation of Acute Appendicitis in Children and Young Adults.

    Science.gov (United States)

    Kinner, Sonja; Pickhardt, Perry J; Riedesel, Erica L; Gill, Kara G; Robbins, Jessica B; Kitchin, Douglas R; Ziemlewicz, Timothy J; Harringa, John B; Reeder, Scott B; Repplinger, Michael D

    2017-10-01

    Appendicitis is frequently diagnosed in the emergency department, most commonly using CT. The purpose of this study was to compare the diagnostic accuracy of contrast-enhanced MRI with that of contrast-enhanced CT for the diagnosis of appendicitis in adolescents when interpreted by abdominal radiologists and pediatric radiologists. Our study included a prospectively enrolled cohort of 48 patients (12-20 years old) with nontraumatic abdominal pain who underwent CT and MRI. Fellowship-trained abdominal and pediatric radiologists reviewed all CT and MRI studies in randomized order, blinded to patient outcome. Likelihood for appendicitis was rated on a 5-point scale (1, definitely not appendicitis; 5, definitely appendicitis) for CT, the unenhanced portion of the MRI, and the entire contrast-enhanced MRI study. ROC curves were generated and AUC compared for each scan type for all six readers and then stratified by radiologist type. Image test characteristics, interrater reliability, and reading times were compared. Sensitivity and specificity were 85.9% (95% CI, 76.2-92.7%) and 93.8% (95% CI, 89.7-96.7%) for unenhanced MRI, 93.6% (95% CI, 85.6-97.9%) and 94.3% (95% CI, 90.2-97%) for contrast-enhanced MRI, and 93.6% (95% CI, 85.6-97.9%) and 94.3% (95% CI, 90.2-97%) for CT. No difference was found in the diagnostic accuracy or interpretation time when comparing abdominal radiologists to pediatric radiologists (CT, 3.0 min vs 2.8 min; contrast-enhanced MRI, 2.4 min vs 1.8 min; unenhanced MRI, 1.5 min vs 2.3 min). Substantial agreement between abdominal and pediatric radiologists was seen for all methods (κ = 0.72-0.83). The diagnostic accuracy of MRI to diagnose appendicitis was very similar to CT. No statistically significant difference in accuracy was observed between imaging modality or radiologist subspecialty.

  5. Diagnostic accuracy of CT, US and angiography in histologically verified hepatocellular carcinoma. Report on 125 foci in 84 cases

    Energy Technology Data Exchange (ETDEWEB)

    Goto, Hiroo; Moriyama, Noriyuki; Shima, Yasuo; Muramatsu, Yukio; Takayasu, Kenichi; Ushio, Kyosuke; Matsue, Hiroto; Sasagawa, Michizo; Yamada, Tatsuya

    1985-02-01

    Detectability of a total of 125 foci in 84 cases with HCC was evaluated. All foci were surgically resected and confirmed histologically. Tumor size >=2.1 cm in diameter:each modality has the accuracy more than 90%, 2.0>=Tumor size>=1.1 cm:CT has the accuracy of 81%, US 56% and Angio 56% respectively. 1.0 cm>=Tumor size:CT shows the accuracy of 35%, US 5% and Angio 30%. Combined with CT, US and Angio in the lesions more than 1.1 cm in diameter, over all accuracy was 100%. Whereas, in the foci less than 1.0 cm, over all accuracy was only 45%, comparing with that of intraoperative US. 57%. Therefore, it was clarified to be very difficult to detect these small lesions, and a careful survey is needed. (author).

  6. Derivation of linear attenuation coefficients from CT numbers for low-energy photons

    International Nuclear Information System (INIS)

    Watanabe, Y.

    1999-01-01

    One can estimate photon attenuation properties from the CT number. In a standard method one assumes that the linear attenuation coefficient is proportional to electron density and ignores its nonlinear dependence on atomic number. When the photon energy is lower than about 50 keV, such as for brachytherapy applications, however, photoelectric absorption and Rayleigh scattering become important. Hence the atomic number must be explicitly considered in estimating the linear attenuation coefficient. In this study we propose a method to more accurately estimate the linear attenuation coefficient of low-energy photons from CT numbers. We formulate an equation that relates the CT number to the electron density and the effective atomic number. We use a CT calibration phantom to determine unknown coefficients in the equation. The equation with a given CT number is then solved for the effective atomic number, which in turn is used to calculate the linear attenuation coefficient for low-energy photons. We use the CT phantom to test the new method. The method significantly improves the standard method in estimating the attenuation coefficient at low photon energies (20keV≤E≤40keV) for materials with high atomic numbers. (author)

  7. Prospective Comparison of the Diagnostic Accuracy of MR Imaging versus CT for Acute Appendicitis.

    Science.gov (United States)

    Repplinger, Michael D; Pickhardt, Perry J; Robbins, Jessica B; Kitchin, Douglas R; Ziemlewicz, Tim J; Hetzel, Scott J; Golden, Sean K; Harringa, John B; Reeder, Scott B

    2018-04-24

    Purpose To compare the accuracy of magnetic resonance (MR) imaging with that of computed tomography (CT) for the diagnosis of acute appendicitis in emergency department (ED) patients. Materials and Methods This was an institutional review board-approved, prospective, observational study of ED patients at an academic medical center (February 2012 to August 2014). Eligible patients were nonpregnant and 12- year-old or older patients in whom a CT study had been ordered for evaluation for appendicitis. After informed consent was obtained, CT and MR imaging (with non-contrast material-enhanced, diffusion-weighted, and intravenous contrast-enhanced sequences) were performed in tandem, and the images were subsequently retrospectively interpreted in random order by three abdominal radiologists who were blinded to the patients' clinical outcomes. Likelihood of appendicitis was rated on a five-point scale for both CT and MR imaging. A composite reference standard of surgical and histopathologic results and clinical follow-up was used, arbitrated by an expert panel of three investigators. Test characteristics were calculated and reported as point estimates with 95% confidence intervals (CIs). Results Analysis included images of 198 patients (114 women [58%]; mean age, 31.6 years ± 14.2 [range, 12-81 years]; prevalence of appendicitis, 32.3%). The sensitivity and specificity were 96.9% (95% CI: 88.2%, 99.5%) and 81.3% (95% CI: 73.5%, 87.3%) for MR imaging and 98.4% (95% CI: 90.5%, 99.9%) and 89.6% (95% CI: 82.8%, 94.0%) for CT, respectively, when a cutoff point of 3 or higher was used. The positive and negative likelihood ratios were 5.2 (95% CI: 3.7, 7.7) and 0.04 (95% CI: 0, 0.11) for MR imaging and 9.4 (95% CI: 5.9, 16.4) and 0.02 (95% CI: 0.00, 0.06) for CT, respectively. Receiver operating characteristic curve analysis demonstrated that the optimal cutoff point to maximize accuracy was 4 or higher, at which point there was no difference between MR imaging and CT

  8. Accuracy verification of PET-CT image fusion and its utilization in target delineation of radiotherapy

    International Nuclear Information System (INIS)

    Wang Xuetao; Yu Jinming; Yang Guoren; Gong Heyi

    2005-01-01

    Objective: Evaluate the accuracy of co-registration of PET and CT (PET-CT) images on line with phantom, and utilize it on patients to provide clinical evidence for target delineation in radiotherapy. Methods: A phantom with markers and different volume cylinders was infused with various concentrations of 18 FDG, and scanned at 4 mm by PET and CT respectively. After having been transmitted into GE eNTEGRA and treatment planning system (TPS) workstations, the images were fused and reconstructed. The distance between the markers and the errors were monitored in PET and CT images respectively. The volume of cylinder in PET and CT images were measured and compared by certain pixel value proportion deduction method. The same procedure was performed on the pulmonary tumor image in ten patients. Results: eNTEGRA and TPS workstations had a good length linearity, but the fusion error of the latter was markedly greater than the former. Tumors in different volume filled by varying concentrations of 18 FDG required different pixel deduction proportion. The cylinder volume of PET and CT images were almost the same, so were the images of pulmonary tumor of ten patients. Conclusions: The accuracy of image co-registration of PET-CT on line may fulfill the clinical demand. Pixel value proportion deduction method can be used for target delineation on PET image. (authors)

  9. Accuracy of lung nodule volumetry in low-dose CT with iterative reconstruction: an anthropomorphic thoracic phantom study.

    Science.gov (United States)

    Doo, K W; Kang, E-Y; Yong, H S; Woo, O H; Lee, K Y; Oh, Y-W

    2014-09-01

    The purpose of this study was to assess accuracy of lung nodule volumetry in low-dose CT with application of iterative reconstruction (IR) according to nodule size, nodule density and CT tube currents, using artificial lung nodules within an anthropomorphic thoracic phantom. Eight artificial nodules (four diameters: 5, 8, 10 and 12 mm; two CT densities: -630 HU that represents ground-glass nodule and +100 HU that represents solid nodule) were randomly placed inside a thoracic phantom. Scans were performed with tube current-time product to 10, 20, 30 and 50 mAs. Images were reconstructed with IR and filtered back projection (FBP). We compared volume estimates to a reference standard and calculated the absolute percentage error (APE). The APE of all nodules was significantly lower when IR was used than with FBP (7.5 ± 4.7% compared with 9.0 ±6.9%; p volumetry in low-dose CT by application of IR showed reliable accuracy in a phantom study. Lung nodule volumetry can be reliably applicable to all lung nodules including small, ground-glass nodules even in ultra-low-dose CT with application of IR. IR significantly improved the accuracy of lung nodule volumetry compared with FBP particularly for ground-glass (-630 HU) nodules. Volumetry in low-dose CT can be utilized in patient with lung nodule work-up, and IR has benefit for small, ground-glass lung nodules in low-dose CT.

  10. Diagnostic accuracy of static CT perfusion for the detection of myocardial ischemia

    DEFF Research Database (Denmark)

    Sørgaard, Mathias Holm; Kofoed, Klaus Fuglsang; Linde, Jesper James

    2016-01-01

    : Systematic literature review and meta-analysis of studies examining the diagnostic accuracy of static CTP imaging alone or combined with coronary CT angiography (CTA) in comparison to single photon emission computed tomography (SPECT), magnetic resonance perfusion (MRP), and/or invasive coronary angiography...

  11. A systematic review on diagnostic accuracy of CT-based detection of significant coronary artery disease

    International Nuclear Information System (INIS)

    Janne d'Othee, Bertrand; Siebert, Uwe; Cury, Ricardo; Jadvar, Hossein; Dunn, Edward J.; Hoffmann, Udo

    2008-01-01

    Objectives: Systematic review of diagnostic accuracy of contrast enhanced coronary computed tomography (CE-CCT). Background: Noninvasive detection of coronary artery stenosis (CAS) by CE-CCT as an alternative to catheter-based coronary angiography (CCA) may improve patient management. Methods: Forty-one articles published between 1997 and 2006 were included that evaluated native coronary arteries for significant stenosis and used CE-CCT as diagnostic test and CCA as reference standard. Study group characteristics, study methodology and diagnostic outcomes were extracted. Pooled summary sensitivity and specificity of CE-CCT were calculated using a random effects model (1) for all coronary segments, (2) assessable segments, and (3) per patient. Results: The 41 studies totaled 2515 patients (75% males; mean age: 59 years, CAS prevalence: 59%). Analysis of all coronary segments yielded a sensitivity of 95% (80%, 89%, 86%, 98% for electron beam CT, 4/8-slice, 16-slice and 64-slice MDCT, respectively) for a specificity of 85% (77%, 84%, 95%, 91%). Analysis limited to segments deemed assessable by CT showed sensitivity of 96% (86%, 85%, 98%, 97%) for a specificity of 95% (90%, 96%, 96%, 96%). Per patient, sensitivity was 99% (90%, 97%, 99%, 98%) and specificity was 76% (59%, 81%, 83%, 92%). Heterogeneity was quantitatively important but not explainable by patient group characteristics or study methodology. Conclusions: Current diagnostic accuracy of CE-CCT is high. Advances in CT technology have resulted in increases in diagnostic accuracy and proportion of assessable coronary segments. However, per patient, accuracy may be lower and CT may have more limited clinical utility in populations at high risk for CAD

  12. Diagnostic Accuracy of the Volume Rendering Images of Multi-Detector CT for the Detection of Lumbar Transverse Process Fractures

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Yun Hak; Chun, Tong Jin [Dept. of Radiology, Eulji University Hospital, Daejeon (Korea, Republic of)

    2012-01-15

    To compare the accuracy of three-dimensional computed tomographic (3D CT) volume rendering techniques with axial images of multi-detector row computed tomography to identify lumbar transverse process (LTP) fractures in trauma patients. We retrospectively evaluated 42 patients with back pain as a result of blunt trauma between January and June of 2010. Two radiologists examined the 3D CT volume rendering images independently. The confirmation of a LTP fracture was based on the consensus of the axial images by the two radiologists. The results of 3D CT volume rendering images were compared with the axial images and the diagnostic powers (sensitivity, specificity, and accuracy) were calculated. Seven of the 42 patients had twenty five lumbar transverse process fractures. The diagnostic power of the 3D CT volume rendering technique is as accurate as axial images. Reader 1, sensitivity 96%, specificity 100%, accuracy 99.9%; and Reader 2 sensitivity 100%, specificity 99.8%, accuracy 99.8%. The accordance of the two radiologists was 99.8%. 3D CT volume rendering images can alternate axial images to detect lumbar transverse process fractures with good image quality.

  13. Diagnostic Accuracy of Detecting Hashimoto's Thyroiditis in Thyroid Cancer Patients Who Underwent Thyroid Surgery: Comparison of Ultrasonography, Positron Emission Tomography/CT, Contrast Enhanced CT, and Anti-Thyroid Antibody

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Young Gyun; Lee, Tae Hyun; Park, Dong Hee; Nam, Sang Been [Dept. of Radiology, Korea Cancer Center Hospital, Seoul (Korea, Republic of)

    2012-11-15

    To compare the diagnostic accuracy of ultrasonography (US), F18-fluorodeoxyglucose positron emission tomography/CT (PET/CT), contrast enhanced CT (CECT), serum anti-thyroid antibody for detecting Hashimoto's thyroiditis in thyroid cancer patients who underwent neck surgery. A total of 150 patients with suspicious for thyroid cancer, who had previously undergone US guided needle aspiration of thyroid, were evaluated with the use of US, PET/CT, CECT and serum anti-thyroid antibody. The four studies were performed within two months before neck surgery. Hashimoto's thyroiditis was confirmed by histopathological results. The diagnostic accuracy of US, PET/CT, CECT and serum anti-thyroid antibody were calculated statistically. Hashimoto's thyroiditis was diagnosed in 51 out of the 150 patients, following neck surgery. The sensitivity, specificity, positive predictive value, negative predictive value and accuracy of US were 76.5%, 92.9%, 84.8%, 88.5%, and 87.3%, respectively. The corresponding values of PET/CT were 37.3%, 96.0%, 82.6%, 74.8%, and 76.0%, and CECT were 62.7%, 89.9%, 76.2%, 82.4%, and 80.7%, and serum anti-thyroid antibody level were 90.2%, 93.9%, 88.5%, 94.9%, and 92.7%, respectively. McNemar test revealed significant difference among PET/CT and others, but no significant differences among US, CECT and serum anti-thyroid antibody. Overall, serum anti-thyroid antibody showed most accurate diagnostic performance. In detecting Hashimoto's thyroiditis, serum anti-thyroid antibody showed higher diagnostic accuracy than others. US also showed relatively high diagnostic accuracy.

  14. Methodological quality of diagnostic accuracy studies on non-invasive coronary CT angiography: influence of QUADAS (Quality Assessment of Diagnostic Accuracy Studies included in systematic reviews) items on sensitivity and specificity

    International Nuclear Information System (INIS)

    Schueler, Sabine; Walther, Stefan; Schuetz, Georg M.; Schlattmann, Peter; Dewey, Marc

    2013-01-01

    To evaluate the methodological quality of diagnostic accuracy studies on coronary computed tomography (CT) angiography using the QUADAS (Quality Assessment of Diagnostic Accuracy Studies included in systematic reviews) tool. Each QUADAS item was individually defined to adapt it to the special requirements of studies on coronary CT angiography. Two independent investigators analysed 118 studies using 12 QUADAS items. Meta-regression and pooled analyses were performed to identify possible effects of methodological quality items on estimates of diagnostic accuracy. The overall methodological quality of coronary CT studies was merely moderate. They fulfilled a median of 7.5 out of 12 items. Only 9 of the 118 studies fulfilled more than 75 % of possible QUADAS items. One QUADAS item (''Uninterpretable Results'') showed a significant influence (P = 0.02) on estimates of diagnostic accuracy with ''no fulfilment'' increasing specificity from 86 to 90 %. Furthermore, pooled analysis revealed that each QUADAS item that is not fulfilled has the potential to change estimates of diagnostic accuracy. The methodological quality of studies investigating the diagnostic accuracy of non-invasive coronary CT is only moderate and was found to affect the sensitivity and specificity. An improvement is highly desirable because good methodology is crucial for adequately assessing imaging technologies. (orig.)

  15. Methodological quality of diagnostic accuracy studies on non-invasive coronary CT angiography: influence of QUADAS (Quality Assessment of Diagnostic Accuracy Studies included in systematic reviews) items on sensitivity and specificity

    Energy Technology Data Exchange (ETDEWEB)

    Schueler, Sabine; Walther, Stefan; Schuetz, Georg M. [Humboldt-Universitaet zu Berlin, Freie Universitaet Berlin, Charite Medical School, Department of Radiology, Berlin (Germany); Schlattmann, Peter [University Hospital of Friedrich Schiller University Jena, Department of Medical Statistics, Informatics, and Documentation, Jena (Germany); Dewey, Marc [Humboldt-Universitaet zu Berlin, Freie Universitaet Berlin, Charite Medical School, Department of Radiology, Berlin (Germany); Charite, Institut fuer Radiologie, Berlin (Germany)

    2013-06-15

    To evaluate the methodological quality of diagnostic accuracy studies on coronary computed tomography (CT) angiography using the QUADAS (Quality Assessment of Diagnostic Accuracy Studies included in systematic reviews) tool. Each QUADAS item was individually defined to adapt it to the special requirements of studies on coronary CT angiography. Two independent investigators analysed 118 studies using 12 QUADAS items. Meta-regression and pooled analyses were performed to identify possible effects of methodological quality items on estimates of diagnostic accuracy. The overall methodological quality of coronary CT studies was merely moderate. They fulfilled a median of 7.5 out of 12 items. Only 9 of the 118 studies fulfilled more than 75 % of possible QUADAS items. One QUADAS item (''Uninterpretable Results'') showed a significant influence (P = 0.02) on estimates of diagnostic accuracy with ''no fulfilment'' increasing specificity from 86 to 90 %. Furthermore, pooled analysis revealed that each QUADAS item that is not fulfilled has the potential to change estimates of diagnostic accuracy. The methodological quality of studies investigating the diagnostic accuracy of non-invasive coronary CT is only moderate and was found to affect the sensitivity and specificity. An improvement is highly desirable because good methodology is crucial for adequately assessing imaging technologies. (orig.)

  16. Stopping power accuracy and achievable spatial resolution of helium ion imaging using a prototype particle CT detector system

    Directory of Open Access Journals (Sweden)

    Volz Lennart

    2017-09-01

    Full Text Available A precise relative stopping power map of the patient is crucial for accurate particle therapy. Charged particle imaging determines the stopping power either tomographically – particle computed tomography (pCT – or by combining prior knowledge from particle radiography (pRad and x-ray CT. Generally, multiple Coulomb scattering limits the spatial resolution. Compared to protons, heavier particles scatter less due to their lower charge/mass ratio. A theoretical framework to predict the most likely trajectory of particles in matter was developed for light ions up to carbon and was found to be the most accurate for helium comparing for fixed initial velocity. To further investigate the potential of helium in particle imaging, helium computed tomography (HeCT and radiography (HeRad were studied at the Heidel-berg Ion-Beam Therapy Centre (HIT using a prototype pCT detector system registering individual particles, originally developed by the U.S. pCT collaboration. Several phantoms were investigated: modules of the Catphan QA phantom for analysis of spatial resolution and achievable stopping power accuracy, a paediatric head phantom (CIRS and a custom-made phantom comprised of animal meat enclosed in a 2 % agarose mixture representing human tissue. The pCT images were reconstructed applying the CARP iterative reconstruction algorithm. The MTF10% was investigated using a sharp edge gradient technique. HeRad provides a spatial resolution above that of protons (MTF1010%=6.07 lp/cm for HeRad versus MTF10%=3.35 lp/cm for proton radiography. For HeCT, the spatial resolution was limited by the number of projections acquired (90 projections for a full scan. The RSP accuracy for all inserts of the Catphan CTP404 module was found to be 2.5% or better and is subject to further optimisation. In conclusion, helium imaging appears to offer higher spatial resolution compared to proton imaging. In future studies, the advantage of helium imaging compared to other

  17. Diagnostic accuracy of 18F-FDG PET/CT for detection of suspected recurrence in patients with oesophageal carcinoma

    International Nuclear Information System (INIS)

    Sharma, Punit; Jain, Sachin; Karunanithi, Sellam; Malhotra, Arun; Bal, Chandrasekhar; Kumar, Rakesh; Pal, Sujoy; Julka, Pramod Kumar; Thulkar, Sanjay

    2014-01-01

    To evaluate the role of 18 F-FDG PET/CT in the detection of recurrence in patients with oesophageal carcinoma, suspected clinically or following conventional investigations. This was a retrospective study. Data from 180 patients (age 56.3 ± 10.4 years; 126 men, 54 women) with histopathologically proven oesophageal carcinoma (squamous cell 115, adenocarcinoma 59, neuroendocrine carcinoma 4, small cell 1, poorly differentiated 1) who had undergone 227 18 F-FDG PET/CT studies for suspected recurrence were analysed. Recurrence was suspected clinically or following conventional investigations. PET/CT images were revaluated by two nuclear medicine physicians in consensus. Findings were grouped into local, nodal and distant recurrence. Results were compared to those from contrast-enhanced (CE) CT when available (109 patients). Clinical/imaging follow-up (minimum 6 months) with histopathology (when available) was taken as the reference standard. Of the 227 18 F-FDG PET/CT studies,166 were positive and 61 were negative for recurrent disease. PET/CT showed local recurrence in 134, nodal recurrence in 115 and distant recurrence in 47, with more than one site of recurrence in 34. The PET/CT findings were true-positive in 153 studies, true-negative in 54, false-positive in 13 and false-negative in 7. The sensitivity of 18 F-FDG PET/CT was 96 %, the specificity was 81 %, the positive and negative predictive values were 92 % and 89 %, respectively, and the accuracy was 91 %. PET/CT showed similar accuracy in patients with squamous cell carcinoma and in those with adenocarcinoma (P = 0.181). 18 F-FDG PET/CT was more specific than CECT (67 % vs. 21 %; P 18 F-FDG PET/CT shows high accuracy in the detection of suspected recurrence in patients with oesophageal carcinoma. It is more specific than and is superior to CECT in the detection of nodal recurrence. (orig.)

  18. Improving Intensity-Based Lung CT Registration Accuracy Utilizing Vascular Information

    Directory of Open Access Journals (Sweden)

    Kunlin Cao

    2012-01-01

    Full Text Available Accurate pulmonary image registration is a challenging problem when the lungs have a deformation with large distance. In this work, we present a nonrigid volumetric registration algorithm to track lung motion between a pair of intrasubject CT images acquired at different inflation levels and introduce a new vesselness similarity cost that improves intensity-only registration. Volumetric CT datasets from six human subjects were used in this study. The performance of four intensity-only registration algorithms was compared with and without adding the vesselness similarity cost function. Matching accuracy was evaluated using landmarks, vessel tree, and fissure planes. The Jacobian determinant of the transformation was used to reveal the deformation pattern of local parenchymal tissue. The average matching error for intensity-only registration methods was on the order of 1 mm at landmarks and 1.5 mm on fissure planes. After adding the vesselness preserving cost function, the landmark and fissure positioning errors decreased approximately by 25% and 30%, respectively. The vesselness cost function effectively helped improve the registration accuracy in regions near thoracic cage and near the diaphragm for all the intensity-only registration algorithms tested and also helped produce more consistent and more reliable patterns of regional tissue deformation.

  19. Construction of a voxel model from CT images with density derived from CT numbers

    International Nuclear Information System (INIS)

    Cheng Mengyun; Zeng Qin; Cao Ruifen; Li Gui; Zheng Huaqing; Huang Shanqing; Song Gang; Wu Yican

    2010-01-01

    The voxel models representing human anatomy have been developed to calculate dose distribution in human body, while the density is the most important physical property of voxel model. Traditionally, when creating the Monte Carlo input files, the average tissue parameters recommended in ICRP report were used to assign each voxel in the existing voxel models. However, as each tissue consists of many voxels in which voxels are different in their densities, the method of assigning average tissue parameters doesn't take account of the voxel's discrepancy, and can't represent human anatomy faithfully. To represent human anatomy more faithfully, a method was implemented to assign each voxel, the density of which was derived from CT number. In order to compare with the traditional method, we have constructed two models from a same cadaver specimen date set. A CT-based pelvic voxel model called Pelvis-CT model, was constructed, the densities of which were derived from the CT numbers. A color photograph-based pelvic voxel model called Pelvis-Photo model, was also constructed, the densities of which were taken from ICRP Publication. The CT images and color photographs were obtained from the same female cadaver specimen. The Pelvis-CT and Pelvis-Photo models were ported into Monte Carlo code MCNP to calculate the conversion coefficients from kerma free-in-air to absorbed dose for external monoenergetic photon beams with energies of 0.1, 1 and 10 MeV under anterior-posterior (AP) geometries. The results were compared with those of given in ICRP74. Differences of up to 50% were observed between conversion coefficients of Pelvis-CT and Pelvis-Photo models, moreover the discrepancies decreased for the photon beams with higher energies. The overall trend of conversion coefficients of the Pelvis-CT model were agreed well with that of ICRP74 data. (author)

  20. Construction of a voxel model from CT images with density derived from CT numbers

    International Nuclear Information System (INIS)

    Cheng Mengyun; Zeng Qin; Cao Ruifen; Li Gui; Zheng Huaqing; Huang Shanqing; Song Gang; Wu Yican

    2011-01-01

    The voxel models representing human anatomy have been developed to calculate dose distribution in human body, while the density and elemental composition are the most important physical properties of voxel model. Usually, when creating the Monte Carlo input files, the average tissue densities recommended in ICRP Publication were used to assign each voxel in the existing voxel models. As each tissue consists of many voxels with different densities, the conventional method of average tissue densities failed to take account of the voxel's discrepancy, and therefore could not represent human anatomy faithfully. To represent human anatomy more faithfully, a method was implemented to assign each voxel, the densities of which were derived from CT number. In order to compare with the traditional method, we constructed two models from the cadaver specimen dataset. A CT-based pelvic voxel model called Pelvis-CT model was constructed, the densities of which were derived from the CT numbers. A color photograph-based pelvic voxel model called Pelvis-Photo model was also constructed, the densities of which were taken from ICRP Publication. The CT images and the color photographs were obtained from the same female cadaver specimen. The Pelvis-CT and Pelvis-Photo models were both ported into Monte Carlo code MCNP to calculate the conversion coefficients from kerma free-in-air to absorbed dose for external monoenergetic photon beams with energies of 0.1, 1 and 10 MeV under anterior-posterior (AP) geometry. The results were compared with those of given in ICRP Publication 74. Differences of up to 50% were observed between conversion coefficients of Pelvis-CT and Pelvis- Photo models, moreover the discrepancies decreased for the photon beams with higher energies. The overall trend of conversion coefficients of the Pelvis-CT model agreed well with that of ICRP Publication 74 data. (author)

  1. Early clinical esophageal adenocarcinoma (cT1): Utility of CT in regional nodal metastasis detection and can the clinical accuracy be improved?

    Energy Technology Data Exchange (ETDEWEB)

    Betancourt Cuellar, Sonia L., E-mail: slbetancourt@mdanderson.org; Sabloff, Bradley, E-mail: bsabloff@mdanderson.org; Carter, Brett W., E-mail: bcarter2@mdanderson.org; Benveniste, Marcelo F., E-mail: mfbenveniste@mdanderson.org; Correa, Arlene M., E-mail: amcorrea@mdanderson.org; Maru, Dipen M., E-mail: dmaru@mdanderson.org; Ajani, Jaffer A., E-mail: jajani@mdanderson.org; Erasmus, Jeremy J., E-mail: jerasmus@mdanderson.org; Hofstetter, Wayne L., E-mail: whofstetter@mdanderson.org

    2017-03-15

    Introduction: Treatment of early esophageal cancer depends on the extent of the primary tumor and presence of regional lymph node metastasis.(RNM). Short axis diameter >10 mm is typically used to detect RNM. However, clinical determination of RNM is inaccurate and can result in inappropriate treatment. Purpose of this study is to evaluate the accuracy of a single linear measurement (short axis > 10 mm) of regional nodes on CT in predicting nodal metastasis, in patients with early esophageal cancer and whether using a mean diameter value (short axis + long axis/2) as well as nodal shape improves cN designation. Methods: CTs of 49 patients with cT1 adenocarcinoma treated with surgical resection alone were reviewed retrospectively. Regional nodes were considered positive for malignancy when round or ovoid and mean size >5 mm adjacent to the primary tumor and >7 mm when not adjacent. Results were compared with pN status after esophagectomy. Results: 18/49 patients had pN+ at resection. Using a single short axis diameter >10 mm on CT, nodal metastasis (cN) was positive in 7/49. Only 1 of these patients was pN+ at resection (sensitivity 5%, specificity 80%, accuracy 53%). Using mean size and morphologic criteria, cN was positive in 28/49. 11 of these patients were pN+ at resection (sensitivity 61%, specificity 45%, accuracy 51%). EUS with limited FNA of regional nodes resulted in 16/49 patients with pN+ being inappropriately designated as cN0. Conclusions: Evaluation of size, shape and location of regional lymph nodes on CT improves the sensitivity of cN determination compared with a short axis measurement alone in patients with cT1 esophageal cancer, although clinical utility is limited.

  2. Diagnostic accuracy of (18F-FDG PET/CT for the detection of peritoneal carcinomatosis of colorectal origins

    Directory of Open Access Journals (Sweden)

    Bullier E

    2013-03-01

    Full Text Available Purpose: to evaluate the diagnostic accuracy of (18 F-FDG positron emission tomography/ computerized tomography (PET/CT for the detection of peritoneal carcinomatosis (PC of colorectal cancer (CRC. Methods: one hundred and forty six eligible patients referred for a PET/CT to evaluate CRC at a single institution were included consecutively and retrospectively. After joint training on 20 similar patient files, two nuclear physicians reviewed the PET/CT scans blindly and independently using a method specifically designed for PC detection. The gold standard was either histological results from surgical exploration for resected patients (n=65 or clinical follow up defined as the 3-month CT follow up supplemented by one year of clinical information and CT monitoring for non-resected patients (n=81. As secondary objectives we evaluated the interobserver reliability between the two PET/CT readings and the diagnostic accuracy of a contemporary ceCT (n=69 or CT component of the PET/CT (n=77 reviewed by an independent radiologist . The extent of PC according to a simplified Sugarbaker index (PCI was examined for the operated subgroup, when PCI was available from the surgeon and PET/CT (n=12. Results: according to the gold standard, 35/146 patients had PC (including 19 of 65 patients with histological gold standard and 16 of 81 patients with clinical follow up gold standard. Sensitivity and specificity of PET/CT were 88% and 96% respectively and positive and negative predictive values and accuracy, 88%, 96% and 94% respectively. For CT alone, the corresponding values were: 68%, 92%, 72%, 90% and 86%. The interobserver agreement for the detection of peritoneal carcinomatosis in PET/CT showed high agreement at 0.91 (Kappa. Median PCI was 3 [range: 1–13] when assessed by surgeons and 4 [range: 1–10] on PET/CT. Focal uptake was the most frequently observed sign on PET. Conclusions: PET/CT appears to be an accurate and reproducible test for PC diagnosis in

  3. Accuracy of computed tomography-intravenous cholangiography (CT-IVC) in detection of choledocholithiasis)

    International Nuclear Information System (INIS)

    Gibson, R.; Speer, A.G.; Collier, N.A.; Vincent, J.M.; Jardine, C.; Noack, K.

    2002-01-01

    Full text: The aim of the study was to determine the accuracy of computed tomography-intravenous cholangiography (CT-IVC) in the detection of choledocholithiasis, compared with endoscopic retrograde cholangiography (ERC). 65 patients undergoing ERC had CT-IVC within 24 hours prior to ERC. Helical CT-IVC was performed following intravenous infusion of 100ml Biliscopin (Schering, Berlin). 2mm beam collimation was used with axial reformats at 1.5mm intervals used for image interpretation. Patients with serum bilirubin levels>3 times normal were excluded. CT-IVC and ERC studies were interpreted independently, and studies were graded as positive, negative or indeterminate for the presence of ductal calculi. 65 patients had a median age of 56 years. Median serum bilirubin at the time of CT-IVC was 13 mmol/L (range 5-47). ERC was indeterminate in 3 patients (4.7%) and CT-IVC in 4 (6.3%). 23 patients had ductal calculi on ERC and CT-IVC was positive in 22 (sensitivity 96%). False positive rate for CT-IVC was 8.3% with positive and negative predictive values of 92% and 97% respectively. Stones' size range was 4-16mm. Of the 14 solitary stones, 10 were pound 5mm and 8 pound 4mm.The bilirubin level in the positive cases was within the normal range in 20, with a range of 7-37 mmol/L. CT-IVC is highly accurate for detection of ductal calculi, including single small calculi, in patients with a serum bilirubin of <3 times normal. Copyright (2002) Blackwell Science Pty Ltd

  4. Accuracy of hepatocellular carcinoma detection on multidetector CT in a transplant liver population with explant liver correlation

    International Nuclear Information System (INIS)

    Addley, H.C.; Griffin, N.; Shaw, A.S.; Mannelli, L.; Parker, R.A.; Aitken, S.; Wood, H.; Davies, S.; Alexander, G.J.; Lomas, D.J.

    2011-01-01

    Aim: To evaluate the diagnostic accuracy of multidetector computed tomography (MDCT) for hepatocellular carcinoma (HCC) in cirrhotic patients undergoing liver transplantation. Secondary aims were to examine the effect of radiologist experience and lesion size on diagnostic accuracy. Materials and methods: Thirty-nine patients (72% male with a mean age of 56.5 years) underwent liver transplantation following preoperative triple-phase MDCT examination of the liver. MDCT examinations were retrospectively independently reviewed by three radiologists for the presence and location of suspected HCCs, with the diagnostic confidence recorded using a five-point confidence scale. MDCT examinations were compared with explant specimens for histopathological correlation. Results: Histopathological results demonstrated 46 HCCs in 29 of the 39 patients. Analysis demonstrated a sensitivity of 65-75% and specificity of 47-88% for detection of HCC lesions. The sensitivity dropped to 48-57% for lesions of size ≤20 mm. As the diagnostic confidence increased, there was a further decrease in the sensitivity (4-26%). The radiologist with the greatest number of years experience was found to have a significantly higher accuracy of detection of HCC lesions compared with the least experienced radiologist. Conclusion: Larger lesion size of HCC and greater number of years experience of the radiologist resulted in significantly higher accuracy of HCC lesion detection. The overall sensitivity and specificity results for MDCT detection of HCC are comparable to previous helical CT imaging.

  5. Diagnostic Accuracy of CT in Paediatric Intracranial Neoplastic Lesions - Radiologic and Pathologic Correlation

    International Nuclear Information System (INIS)

    Qureshi, A.

    2011-01-01

    The frequency of paediatric tumours in developing countries could be attributed to the increased percentage (39% of total population of children) in the overall population. Therefore, extensive researches should be under taken in the field of Paediatric Oncology in the third world. Objective: This study was conducted to determine the diagnostic accuracy of CT by comparing the pre-operative radiological findings of paediatric brain tumours with post-operative histopathological findings on the basis of characteristic radiological features of various tumours. Materials and Methods: This was a hospital based prospective, cross-sectional and descriptive study carried out in Radiology Dept, KEMU / Mayo Hospital, Lahore. Study was conducted over a period of 3 years from June 2005 till June 2008 and comprised of 100 cases of paediatric brain tumours up to 12 years of age. Cases were also collected from Mayo and Children Hospital, Lahore. Results: Topographically, supratentorial tumours were found more than infratentorial 55 : 45. Low grade were more common than high grade 73 : 27. The most common tumour was astrocytoma with 52 cases. Medulloblastoma ranked the second with 16 cases followed by craniopharyngioma with 12 cases. Conclusion: The diagnostic accuracy of CT scan was found to be 83% when correlated with histopathology. CT proved fairly accurate in detection of paediatric intracranial neoplastic lesions. As CT is relatively commonly available inexpensive modality than MRI so it can be used as non invasive imaging modality. (author)

  6. Diagnostic accuracy of CT in assessing extra-regional lymphadenopathy in pancreatic and peri-ampullary cancer: a systematic review and meta-analysis.

    Science.gov (United States)

    Tseng, Dorine S J; van Santvoort, Hjalmar C; Fegrachi, Samira; Besselink, Marc G; Zuithoff, Nicolaas P A; Borel Rinkes, Inne H; van Leeuwen, Maarten S; Molenaar, I Quintus

    2014-12-01

    Computed tomography (CT) is the most widely used method to assess resectability of pancreatic and peri-ampullary cancer. One of the contra-indications for curative resection is the presence of extra-regional lymph node metastases. This meta-analysis investigates the accuracy of CT in assessing extra-regional lymph node metastases in pancreatic and peri-ampullary cancer. We systematically reviewed the literature according to the PRISMA guidelines. Studies reporting on CT assessment of extra-regional lymph nodes in patients undergoing pancreatoduodenectomy were included. Data on baseline characteristics, CT-investigations and histopathological outcomes were extracted. Diagnostic accuracy, positive predictive value (PPV), negative predictive value (NPV), sensitivity and specificity were calculated for individual studies and pooled data. After screening, 4 cohort studies reporting on CT-findings and histopathological outcome in 157 patients with pancreatic or peri-ampullary cancer were included. Overall, diagnostic accuracy, specificity and NPV varied from 63 to 81, 80-100% and 67-90% respectively. However, PPV and sensitivity ranged from 0 to 100% and 0-38%. Pooled sensitivity, specificity, PPV and NPV were 25%, 86%, 28% and 84% respectively. CT has a low diagnostic accuracy in assessing extra-regional lymph node metastases in pancreatic and peri-ampullary cancer. Therefore, suspicion of extra-regional lymph node metastases on CT alone should not be considered a contra-indication for exploration. Copyright © 2014 Elsevier Ltd. All rights reserved.

  7. The registration accuracy analysis of different CT-MRI imaging fusion method in brain tumor

    International Nuclear Information System (INIS)

    Lu Jie; Yin Yong; Shao Qian; Zhang Zicheng; Chen Jinhu; Chen Zhaoqiu

    2010-01-01

    Objective: To find an effective CT-MRI image fusion protocol in brain tumor by analyzing the registration accuracy of different methods. Methods: The simulation CT scan and MRI T 1 WI imaging of 10 brain tumor patients obtained with same position were registered by Tris-Axes landmark ,Tris-Axes landmark + manual adjustment, mutual information and mutual information + manual adjustment method. The clinical tumor volume (CTV) were contoured on both CT and MRI images respectively. The accuracy of image fusion was assessed by the mean distance of five bone markers (d 1-5 ), central position of CTV (d CTV ) the percentage of CTV overlap (P CT-MRI ) between CT and MRI images. The difference between different methods was analyzed by Freedman M non-parameter test. Results: The difference of the means d1-5 between the Tris-Axes landmark,Tris-Axes landmark plus manual adjustment,mutual information and mutual information plus manual adjustment methods were 0.28 cm ±0.12 cm, 0.15 cm ±0.02 cm, 0.25 cm± 0.19 cm, 0.10 cm ± 0.06 cm, (M = 14.41, P = 0.002). the means d CTV were 0.59 cm ± 0.28 cm, 0.60 cm± 0.32 cm, 0.58 cm ± 0.39 cm, 0.42 cm± 0.30 cm (M = 9.72, P = 0.021), the means P CT-MRI were 0.69% ±0.18%, 0.68% ±0.16%, 0.66% ±0.17%, 0.74% ±0.14% (M =14.82, P=0.002), respectively. Conclusions: Mutual information plus manual adjustment registration method was the preferable fusion method for brain tumor patients. (authors)

  8. Diagnostic accuracy of {sup 18}F-FDG PET/CT compared with that of contrast-enhanced MRI of the breast at 3 T

    Energy Technology Data Exchange (ETDEWEB)

    Magometschnigg, Heinrich F.; Baltzer, Pascal A.; Fueger, Barbara; Helbich, Thomas H.; Weber, Michael [Medical University of Vienna, Department of Biomedical Imaging and Image-guided Therapy, Division of Molecular and Gender Imaging, Vienna (Austria); Karanikas, Georgios [Medical University of Vienna, Department of Biomedical Imaging and Image-guided Therapy, Division of Nuclear Medicine, Vienna (Austria); Dubsky, Peter [Medical University of Vienna, Department of Surgery, Vienna (Austria); Rudas, Margaretha [Medical University of Vienna, Department of Pathology, Vienna (Austria); Pinker, Katja [Medical University of Vienna, Department of Biomedical Imaging and Image-guided Therapy, Division of Molecular and Gender Imaging, Vienna (Austria); Molecular Imaging and Therapy Service, Memorial Sloan Kettering Cancer Center, New York (United States)

    2015-10-15

    To compare the diagnostic accuracy of prone {sup 18}F-FDG PET/CT with that of contrast-enhanced MRI (CE-MRI) at 3 T in suspicious breast lesions. To evaluate the influence of tumour size on diagnostic accuracy and the use of maximum standardized uptake value (SUV{sub MAX}) thresholds to differentiate malignant from benign breast lesions. A total of 172 consecutive patients with an imaging abnormality were included in this IRB-approved prospective study. All patients underwent {sup 18}F-FDG PET/CT and CE-MRI of the breast at 3 T in the prone position. Two reader teams independently evaluated the likelihood of malignancy as determined by {sup 18}F-FDG PET/CT and CE-MRI independently. {sup 18}F-FDG PET/CT data were qualitatively evaluated by visual interpretation. Quantitative assessment was performed by calculation of SUV{sub MAX}. Sensitivity, specificity, diagnostic accuracy, area under the curve and interreader agreement were calculated for all lesions and for lesions <10 mm. Histopathology was used as the standard of reference. There were 132 malignant and 40 benign lesions; 23 lesions (13.4 %) were <10 mm. Both {sup 18}F-FDG PET/CT and CE-MRI achieved an overall diagnostic accuracy of 93 %. There were no significant differences in sensitivity (p = 0.125), specificity (p = 0.344) or diagnostic accuracy (p = 1). For lesions <10 mm, diagnostic accuracy deteriorated to 91 % with both {sup 18}F-FDG PET/CT and CE-MRI. Although no significant difference was found for lesions <10 mm, CE-MRI at 3 T seemed to be more sensitive but less specific than {sup 18}F-FDG PET/CT. Interreader agreement was excellent (κ = 0.85 and κ = 0.92). SUV{sub MAX} threshold was not helpful in differentiating benign from malignant lesions. {sup 18}F-FDG PET/CT and CE-MRI at 3 T showed equal diagnostic accuracies in breast cancer diagnosis. For lesions <10 mm, diagnostic accuracy deteriorated, but was equal for {sup 18}F-FDG PET/CT and CE-MRI at 3 T. For lesions <10 mm, CE-MRI at 3 T seemed

  9. Influence of CT on the number of radiologic procedures and hospitalization

    International Nuclear Information System (INIS)

    Binkhuysen, F.H.B.; Puijlaert, C.B.A.J.

    1987-01-01

    The research concerns the influence of body CT on medical efficiency in radiology and hospitalization in The Netherlands. General hospitals with body CT are compared with hospitals without CT. In radiology the substitution effect is investigated with use of the number of radiological performances per clinical patient as a parameter. From 1977 to 1984 this parameter proves to decrease in hospitals with body CT (9%) in contrast to the rise in hospitals without CT (10%). Concerning the average hospital in patient stay, it appears that this parameter shows a faster decrease in hospitals with body CT compared with hospitals without CT (5%-8%)

  10. Experimental estimation of regional lung water volume by histogram of pulmonary CT numbers

    International Nuclear Information System (INIS)

    Kato, Shiro; Momoki, Shigeru; Asai, Toshihiko; Shimada, Takeshi; Tamano, Masahiro; Nakamoto, Takaaki; Yoshimura, Masaharu

    1989-01-01

    Both in vitro and in vivo experiments were made to assess the ability of pulmonary CT numbers to quantitatively determine regional water volume in cases of pulmonary congestion or edema associated with left heart failure. In vitro experiment revealed a good linear correlation between the volume of injected water and the determined CT number of polyethylene tube packed with sponge. In the subsequent in vivo experiment with 10 adult mongrel dogs, lung water volumes obtained by pulmonary CT numbers were found to be consistent with the actual volumes. Pulmonary CT numbers for water volume proved to become parameters to quantitatively evaluate pulmonary congestion or edema. (Namekawa, K)

  11. CT in neuromuscular disorders: A comparison of CT and histology

    International Nuclear Information System (INIS)

    Vliet, A.M. van der; Thijssen, H.O.M.; Merx, J.L.; Joosten, E.

    1988-01-01

    The value of CT-examination of the muscles compared to histology was studied in a retrospective analysis of 30 patients with clinical suspicion of neuromuscular disorder. In the evaluation of the CT-results descriptive criteria were used. The histologic diagnosis came from needle-biopsies taken from the quadriceps muscle. Considering the whole group of neuromuscular disorders, CT has an overall accuracy of 84.8%, a positive predictive value of 95.5% and a negative predictive value of 63.6%. This makes the use of CT as a diagnostic tool in neuromuscular disorders a reliable examination technique. In patients with a polymyositis there is even a 100% correlation between CT findings and biopsy results. Discrepancy between the biopsy results is remarkable of the quadriceps muscle and the CT findings: The number of abnormal histological findings is twice the number of abnormal CT findings. Using the more proximal gluteal region as a biopsy site would have decreased this discrepancy and would therefore have given a better correlation between CT and histology. The choice of protocol in determining the levels to be scanned is of great importance in achieving good reproducability in follow-up CT examinations. (orig.)

  12. Adrenal adenomas: relationship between histologic lipid-rich cells and CT attenuation number

    International Nuclear Information System (INIS)

    Yamada, Takayuki; Ishibashi, Tadashi; Saito, Haruo; Matsuhashi, Toshio; Majima, Kazuhiro; Tsuda, Masashi; Takahashi, Shoki; Moriya, Takuya

    2003-01-01

    Objective: To evaluate the relationship between lipid-rich cells of the adrenal adenoma and precontrast computed tomographic (CT) attenuation numbers in three clinical groups. Materials and Methods: Thirty-five surgically resected adrenal adenomas were used. The clinical diagnoses of the patients included 13 cases of primary aldosteronism, 15 cases of Cushing's syndrome, and 7 non-functioning tumors. The number of lipid-rich clear cells was counted using a microscopic eyepiece grid that contained 100 squares. The results were expressed as the percentages of lipid-rich areas. Results: There was a strong inverse linear relationship between the percentage of lipid-rich cells and the precontrast CT attenuation number (R 2 =0.724, P<0.0001). There were significantly more lipid-rich cells in the primary aldosteronism and non-functioning tumor cases compared to cases of Cushing's syndrome (P=0.007 and 0.015, respectively). The CT attenuation numbers of the primary aldosteronism cases were significantly lower than those of Cushing's syndrome (P=0.0052). Furthermore, the CT attenuation numbers of the non-functioning tumor cases were lower than those of Cushing's syndrome cases. Conclusion: We showed that adrenal adenomas in primary aldosteronism and non-functioning tumors contain significantly more lipid-rich cells than those in Cushing's syndrome. They also showed significantly lower attenuation than that in Cushing's syndrome on CT scans. Our results suggest that precontrast CT attenuation numbers may be helpful in the differentiation of adenomas from non-adenomatous lesions, which include malignancies

  13. Accuracy of low dose CT in the diagnosis of appendicitis in childhood and comparison with USG and standard dose CT.

    Science.gov (United States)

    Yi, Dae Yong; Lee, Kyung Hoon; Park, Sung Bin; Kim, Jee Taek; Lee, Na Mi; Kim, Hyery; Yun, Sin Weon; Chae, Soo Ahn; Lim, In Seok

    Computed tomography should be performed after careful consideration due to radiation hazard, which is why interest in low dose CT has increased recently in acute appendicitis. Previous studies have been performed in adult and adolescents populations, but no studies have reported on the efficacy of using low-dose CT in children younger than 10 years. Patients (n=475) younger than 10 years who were examined for acute appendicitis were recruited. Subjects were divided into three groups according to the examinations performed: low-dose CT, ultrasonography, and standard-dose CT. Subjects were categorized according to age and body mass index (BMI). Low-dose CT was a contributive tool in diagnosing appendicitis, and it was an adequate method, when compared with ultrasonography and standard-dose CT in terms of sensitivity (95.5% vs. 95.0% and 94.5%, p=0.794), specificity (94.9% vs. 80.0% and 98.8%, p=0.024), positive-predictive value (96.4% vs. 92.7% and 97.2%, p=0.019), and negative-predictive value (93.7% vs. 85.7% and 91.3%, p=0.890). Low-dose CT accurately diagnosed patients with a perforated appendix. Acute appendicitis was effectively diagnosed using low-dose CT in both early and middle childhood. BMI did not influence the accuracy of detecting acute appendicitis on low-dose CT. Low-dose CT is effective and accurate for diagnosing acute appendicitis in childhood, as well as in adolescents and young adults. Additionally, low-dose CT was relatively accurate, irrespective of age or BMI, for detecting acute appendicitis. Therefore, low-dose CT is recommended for assessing children with suspected acute appendicitis. Copyright © 2017. Published by Elsevier Editora Ltda.

  14. Acute appendicitis: sensitivity, specificity and diagnostic accuracy of thin-section contrast-enhanced CT findings

    International Nuclear Information System (INIS)

    Lee, Ji Yon; Choi, Dong Il; Park, Hae Won; Lee, Young Rae; Kook, Shin Ho; Kwang, Hyon Joo; Kim, Seung Kwon; Chung, Eun Chul

    2002-01-01

    To assess the sensitivity, specificity, and diagnostic accuracy of individual contrast-enhanced helical CT findings of acute appendicitis. We retrospectively reviewed the appendiceal helical CT scans, obtained after intravenous contrast administration (abdomen; 7-mm collimation, abdominopelvic junction; 5-mm collimation), of 50 patients with surgically proven acute appendicitis and 112 with alternative diagnoses. The following parameters were analysed by three radiologists: enlarged appendix (>6 mm in diameter), appendiceal wall thickening, appendiceal wall enhancement, no identification of the appendix, appendicolith(s), (appendiceal) intraluminal air, abscess, lymphadenopathy, terminal ileal wall thickening, focal cecal apical thickening, focal colonic wall thickening, and segmental colonic wall thickening. The CT findings of acute appendicitis that statistically distinguished it from alternative diagnoses were an enlarged appendix (sensitivity; 92%, specificity; 93%, diagnostic accuracy; 93%), appendiceal wall thickening (for these three parameters: 68%, 96% and 88%, respectively), periappendiceal fat stranding (90%, 79%, 82%), appendiceal wall enhancement (72%, 86%, 82%), appendicolith (16%, 100%, 74%), and focal cecal apical thickening (14%, 100%, 74%) (for each, p<0305). On thin-section contrast-enhanced helical CT, an enlarged appendix and periappendiceal fat stranding were found in 90% or more patients with acute appendicitis. Appendiceal wall thickening and enhancement were alearly demonstrated and significant findings for diagnosis. Less common but specific findings include appendicolith, focal cecal apical thickening and intramural air, can also help us establish a diagnosis of acute appendicitis

  15. Bicuspid aortic valves: Diagnostic accuracy of standard axial 64-slice chest CT compared to aortic valve image plane ECG-gated cardiac CT

    Energy Technology Data Exchange (ETDEWEB)

    Murphy, David J., E-mail: david.murphy@st-vincents.ie [Department of Radiology, St Vincent' s University Hospital, Elm Park, Dublin 4 (Ireland); McEvoy, Sinead H., E-mail: s.mcevoy@st-vincents.ie [Department of Radiology, St Vincent' s University Hospital, Elm Park, Dublin 4 (Ireland); Iyengar, Sri, E-mail: sri.iyengar@nhs.net [Department of Radiology, Plymouth Hospitals NHS Trust, Plymouth Devon PL6 8DH (United Kingdom); Feuchtner, Gudrun, E-mail: Gudrun.Feuchtner@i-med.ac.at [Department of Radiology, Innsbruck Medical University, Anichstr. 35, A-6020 Innsbruck (Austria); Cury, Ricardo C., E-mail: r.cury@baptisthealth.net [Department of Radiology, Baptist Cardiac and Vascular Institute, 8900 North Kendall Drive, Miami, FL 33176 (United States); Roobottom, Carl, E-mail: carl.roobottom@nhs.net [Department of Radiology, Plymouth Hospitals NHS Trust, Plymouth Devon PL6 8DH (United Kingdom); Plymouth University Peninsula Schools of Medicine and Dentistry (United Kingdom); Baumueller, Stephan, E-mail: Hatem.Alkadhi@usz.ch [Institute for Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich (Switzerland); Alkadhi, Hatem, E-mail: stephan.baumueller@usz.ch [Institute for Diagnostic and Interventional Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich (Switzerland); Dodd, Jonathan D., E-mail: jonniedodd@gmail.com [Department of Radiology, St Vincent' s University Hospital, Elm Park, Dublin 4 (Ireland)

    2014-08-15

    Objectives: To assess the diagnostic accuracy of standard axial 64-slice chest CT compared to aortic valve image plane ECG-gated cardiac CT for bicuspid aortic valves. Materials and methods: The standard axial chest CT scans of 20 patients with known bicuspid aortic valves were blindly, randomly analyzed for (i) the appearance of the valve cusps, (ii) the largest aortic sinus area, (iii) the longest aortic cusp length, (iv) the thickest aortic valve cusp and (v) valve calcification. A second blinded reader independently analyzed the appearance of the valve cusps. Forty-two age- and sex-matched patients with known tricuspid aortic valves were used as controls. Retrospectively ECG-gated cardiac CT multiphase reconstructions of the aortic valve were used as the gold-standard. Results: Fourteen (21%) scans were scored as unevaluable (7 bicuspid, 7 tricuspid). Of the remainder, there were 13 evaluable bicuspid valves, ten of which showed an aortic valve line sign, while the remaining three showed a normal Mercedes-Benz appearance owing to fused valve cusps. The 35 evaluable tricuspid aortic valves all showed a normal Mercedes-Benz appearance (P = 0.001). Kappa analysis = 0.62 indicating good interobserver agreement for the aortic valve cusp appearance. Aortic sinus areas, aortic cusp lengths and aortic cusp thicknesses of ≥3.8 cm{sup 2}, 3.2 cm and 1.6 mm respectively on standard axial chest CT best distinguished bicuspid from tricuspid aortic valves (P < 0.0001 for all). Of evaluable scans, the sensitivity, specificity, positive and negative predictive values of standard axial chest CT in diagnosing bicuspid aortic valves was 77% (CI 0.54–1.0), 100%, 100% and 70% respectively. Conclusion: The aortic valve is evaluable in approximately 80% of standard chest 64-slice CT scans. Bicuspid aortic valves may be diagnosed on evaluable scans with good diagnostic accuracy. An aortic valve line sign, enlarged aortic sinuses and elongated, thickened valve cusps are specific CT

  16. Reappraisal of Pediatric Diastatic Skull Fractures in the 3-Dimensional CT Era: Clinical Characteristics and Comparison of Diagnostic Accuracy of Simple Skull X-Ray, 2-Dimensional CT, and 3-Dimensional CT.

    Science.gov (United States)

    Sim, Sook Young; Kim, Hyun Gi; Yoon, Soo Han; Choi, Jong Wook; Cho, Sung Min; Choi, Mi Sun

    2017-12-01

    Diastatic skull fractures (DSFs) in children are difficult to detect in skull radiographs before they develop into growing skull fractures; therefore, little information is available on this topic. However, recent advances in 3-dimensional (3D) computed tomography (CT) imaging technology have enabled more accurate diagnoses of almost all forms of skull fracture. The present study was undertaken to document the clinical characteristics of DSFs in children and to determine whether 3D CT enhances diagnostic accuracy. Two hundred and ninety-two children younger than 12 years with skull fractures underwent simple skull radiography, 2-dimensional (2D) CT, and 3DCT. Results were compared with respect to fracture type, location, associated lesions, and accuracy of diagnosis. DSFs were diagnosed in 44 (15.7%) of children with skull fractures. Twenty-two patients had DSFs only, and the other 22 had DSFs combined with compound or mixed skull fractures. The most common fracture locations were the occipitomastoid (25%) and lambdoid (15.9%). Accompanying lesions consisted of subgaleal hemorrhages (42/44), epidural hemorrhages (32/44), pneumocephalus (17/44), and subdural hemorrhages (3/44). A total of 17 surgical procedures were performed on 15 of the 44 patients. Fourteen and 19 patients were confirmed to have DSFs by skull radiography and 2D CT, respectively, but 3D CT detected DSFs in 43 of the 44 children (P skull radiography or 2D CT for detecting DSFs. This finding indicates that 3D CT should be used routinely rather than 2D CT for the assessment of pediatric head trauma. Copyright © 2017 Elsevier Inc. All rights reserved.

  17. Diagnostic accuracy of {sup 18}F-FDG PET/CT for detection of suspected recurrence in patients with oesophageal carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Sharma, Punit; Jain, Sachin; Karunanithi, Sellam; Malhotra, Arun; Bal, Chandrasekhar; Kumar, Rakesh [All India Institute of Medical Sciences, Department of Nuclear Medicine, New Delhi (India); Pal, Sujoy [All India Institute of Medical Sciences, Department of Surgical Gastroenterology, New Delhi (India); Julka, Pramod Kumar [All India Institute of Medical Sciences, Department of Radiation Oncology, New Delhi (India); Thulkar, Sanjay [All India Institute of Medical Sciences, Department of Radiodiagnosis, New Delhi (India)

    2014-06-15

    To evaluate the role of {sup 18}F-FDG PET/CT in the detection of recurrence in patients with oesophageal carcinoma, suspected clinically or following conventional investigations. This was a retrospective study. Data from 180 patients (age 56.3 ± 10.4 years; 126 men, 54 women) with histopathologically proven oesophageal carcinoma (squamous cell 115, adenocarcinoma 59, neuroendocrine carcinoma 4, small cell 1, poorly differentiated 1) who had undergone 227 {sup 18}F-FDG PET/CT studies for suspected recurrence were analysed. Recurrence was suspected clinically or following conventional investigations. PET/CT images were revaluated by two nuclear medicine physicians in consensus. Findings were grouped into local, nodal and distant recurrence. Results were compared to those from contrast-enhanced (CE) CT when available (109 patients). Clinical/imaging follow-up (minimum 6 months) with histopathology (when available) was taken as the reference standard. Of the 227 {sup 18}F-FDG PET/CT studies,166 were positive and 61 were negative for recurrent disease. PET/CT showed local recurrence in 134, nodal recurrence in 115 and distant recurrence in 47, with more than one site of recurrence in 34. The PET/CT findings were true-positive in 153 studies, true-negative in 54, false-positive in 13 and false-negative in 7. The sensitivity of {sup 18}F-FDG PET/CT was 96 %, the specificity was 81 %, the positive and negative predictive values were 92 % and 89 %, respectively, and the accuracy was 91 %. PET/CT showed similar accuracy in patients with squamous cell carcinoma and in those with adenocarcinoma (P = 0.181).{sup 18}F-FDG PET/CT was more specific than CECT (67 % vs. 21 %; P < 0.0001). PET/CT was superior to CECT for the detection of nodal recurrence (P < 0.0001), but not local recurrence (P = 0.093) or distant metastases (P = 0.441). {sup 18}F-FDG PET/CT shows high accuracy in the detection of suspected recurrence in patients with oesophageal carcinoma. It is more specific than

  18. Performance evaluation of the General Electric eXplore CT 120 micro-CT using the vmCT phantom

    Energy Technology Data Exchange (ETDEWEB)

    Bahri, M.A., E-mail: M.Bahri@ulg.ac.be [ULg-Liege University, Cyclotron Research Centre, Liege, Bat. 30, Allee du 6 aout, 8 (Belgium); Warnock, G.; Plenevaux, A. [ULg-Liege University, Cyclotron Research Centre, Liege, Bat. 30, Allee du 6 aout, 8 (Belgium); Choquet, P.; Constantinesco, A. [Biophysique et Medecine Nucleaire, Hopitaux universitaires de Strasbourg, Strasbourg (France); Salmon, E.; Luxen, A. [ULg-Liege University, Cyclotron Research Centre, Liege, Bat. 30, Allee du 6 aout, 8 (Belgium); Seret, A. [ULg-Liege University, Cyclotron Research Centre, Liege, Bat. 30, Allee du 6 aout, 8 (Belgium); ULg-Liege University, Experimental Medical Imaging, Liege (Belgium)

    2011-08-21

    The eXplore CT 120 is the latest generation micro-CT from General Electric. It is equipped with a high-power tube and a flat-panel detector. It allows high resolution and high contrast fast CT scanning of small animals. The aim of this study was to compare the performance of the eXplore CT 120 with that of the eXplore Ultra, its predecessor for which the methodology using the vmCT phantom has already been described . The phantom was imaged using typical a rat (fast scan or F) or mouse (in vivo bone scan or H) scanning protocols. With the slanted edge method, a 10% modulation transfer function (MTF) was observed at 4.4 (F) and 3.9-4.4 (H) mm{sup -1} corresponding to 114 {mu}m resolution. A fairly larger MTF was obtained by the coil method with the MTF for the thinnest coil (3.3 mm{sup -1}) equal to 0.32 (F) and 0.34 (H). The geometric accuracy was better than 0.3%. There was a highly linear (R{sup 2}>0.999) relationship between measured and expected CT numbers for both the CT number accuracy and linearity sections of the phantom. A cupping effect was clearly seen on the uniform slices and the uniformity-to-noise ratio ranged from 0.52 (F) to 0.89 (H). The air CT number depended on the amount of polycarbonate surrounding the area where it was measured; a difference as high as approximately 200 HU was observed. This hindered the calibration of this scanner in HU. This is likely due to the absence of corrections for beam hardening and scatter in the reconstruction software. However in view of the high linearity of the system, the implementation of these corrections would allow a good quality calibration of the scanner in HU. In conclusion, the eXplore CT 120 achieved a better spatial resolution than the eXplore Ultra (based on previously reported specifications) and future software developments will include beam hardening and scatter corrections that will make the new generation CT scanner even more promising.

  19. Reduction of Cone-Beam CT scan time without compromising the accuracy of the image registration in IGRT

    International Nuclear Information System (INIS)

    Westberg, Jonas; Jensen, Henrik R.; Bertelsen, Anders; Brink, Carsten

    2010-01-01

    Background. In modern radiotherapy accelerators are equipped with 3D cone-beam CT (CBCT) which is used to verify patient position before treatment. The verification is based on an image registration between the CBCT acquired just before treatment and the CT scan made for the treatment planning. The purpose of this study is to minimise the scan time of the CBCT without compromising the accuracy of the image registration in IGRT. Material and methods. Fast scans were simulated by reducing the number of acquired projection images, i.e. new reconstructions based on a subset of the original projections were made. The deviation between the registrations of these new reconstructions and the original registration was measured as function of the amount of reduction. Results and Discussion. Twenty nine head and neck (HandN) and 11 stereotactic lung patients were included in the study. The mean of the registration deviation did not differ significantly from zero independently of the number of projections included in the reconstruction. Except for the smallest subset of reconstructions (10% and 25% of the original projection for the lung and HandN patients, respectively) the standard deviation of the registration differences was constant. The standard deviations were approximately 0.1 mm and 0.2 mm for the HandN and lung group, respectively. Based on these results an in-house developed solution, able to reduce the Cone-Beam CT scan time, has been implemented clinically

  20. Diagnostic accuracy of FDG PET/CT in mediastinal lymph nodal staging of the non-small cell lung cancer: prospective study with PET/CT of 182 cases

    International Nuclear Information System (INIS)

    Lee, J. W.; Kang, W. J.; Kim, B. S.; Lee, D. S.; Jeong, J. K.; Lee, M. C.

    2007-01-01

    This study was performed to assess the accuracy of fluorodeoxyglucose - positron emission tomography/computed tomography (FDG-PET/CT) in the mediastinal lymph nodal staging of non-small cell lung cancer as compared with CT. Between March 2004 and February 2006, 182 patients (126 men and 56 women; mean age, 60.7 y) with non-small cell lung cancer underwent FDG PET/CT and enhanced chest CT. PET/CT and CT images were acquired in a prospective manner. These images were evaluated separately by 2 different physicians and nodal stages were determined by using American Joint Committee on Cancer staging systems. The maxSUV, location, size, calcification and pattern of FDG uptake of lymph nodes were considered. Surgical and histological findings served as the reference standard. A total of 182 patients with 778 mediastinal nodal stations were evaluated. Among them, metastases were found in 36 patients with 53 nodal stations. The respective values for sensitivity, specificity, positive predictive value, negative predictive value and accuracy of mediastinal lymph node staging were 36%, 80%, 30%, 84% and 71% with CT and 75%, 89%, 63%, 94% and 86% with PET/CT on per-patient basis, and 23%, 92%, 18%, 94% and 88% with CT and 66%, 96%, 54%, 98% and 94% with PET/CT on per-nodal-station basis. The maxSUVs of metastatic lymph nodes were significantly higher than those of benign nodes (p = 0.0008). Seventy seven percent (27/35) of the metastatic lymph nodes on FDG-PET/CT images were less than a 1cm in the short axis. Moreover, some benign lymph node patterns, such as bilateral symmetric nodes with similar FDG uptake, benign pattern of nodal calcification and small-sized lymph node with much higher maxSUV than primary tumor, were noted during the evaluation of FDG-PET/CT images. This prospective study suggests that FDG-PET/CT is more accurately stage the mediastinal lymph node staging than CT, and that it provides high specificity and a negative predictive value

  1. Tin-filter enhanced dual-energy-CT: image quality and accuracy of CT numbers in virtual noncontrast imaging.

    Science.gov (United States)

    Kaufmann, Sascha; Sauter, Alexander; Spira, Daniel; Gatidis, Sergios; Ketelsen, Dominik; Heuschmid, Martin; Claussen, Claus D; Thomas, Christoph

    2013-05-01

    To measure and compare the objective image quality of true noncontrast (TNC) images with virtual noncontrast (VNC) images acquired by tin-filter-enhanced, dual-source, dual-energy computed tomography (DECT) of upper abdomen. Sixty-three patients received unenhanced abdominal CT and enhanced abdominal DECT (100/140 kV with tin filter) in portal-venous phase. VNC images were calculated from the DECT datasets using commercially available software. The mean attenuation of relevant tissues and image quality were compared between the TNC and VNC images. Image quality was rated objectively by measuring image noise and the sharpness of object edges using custom-designed software. Measurements were compared using Student two-tailed t-test. Correlation coefficients for tissue attenuation measurements between TNC and VNC were calculated and the relative deviations were illustrated using Bland-Altman plots. Mean attenuation differences between TNC and VNC (HUTNC - HUVNC) image sets were as follows: right liver lobe -4.94 Hounsfield units (HU), left liver lobe -3.29 HU, vena cava -2.19 HU, spleen -7.46 HU, pancreas 1.29 HU, fat -11.14 HU, aorta 1.29 HU, bone marrow 36.83 HU (all P VNC and TNC series were observed for liver, vena portae, kidneys, pancreas, muscle and bone marrow (Pearson's correlation coefficient ≥0.75). Mean image noise was significantly higher in TNC images (P VNC and TNC images (P = .19). The Hounsfield units in VNC images closely resemble TNC images in the majority of the organs of the upper abdomen (kidneys, liver, pancreas). In spleen and fat, Hounsfield numbers in VNC images are tend to be higher than in TNC images. VNC images show a low image noise and satisfactory edge sharpness. Other criteria of image quality and the depiction of certain lesions need to be evaluated additionally. Copyright © 2013 AUR. Published by Elsevier Inc. All rights reserved.

  2. Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT.

    Science.gov (United States)

    Pelgrim, Gert Jan; van Hamersvelt, Robbert W; Willemink, Martin J; Schmidt, Bernhard T; Flohr, Thomas; Schilham, Arnold; Milles, Julien; Oudkerk, Matthijs; Leiner, Tim; Vliegenthart, Rozemarijn

    2017-09-01

    To determine the accuracy of iodine quantification with dual energy computed tomography (DECT) in two high-end CT systems with different spectral imaging techniques. Five tubes with different iodine concentrations (0, 5, 10, 15, 20 mg/ml) were analysed in an anthropomorphic thoracic phantom. Adding two phantom rings simulated increased patient size. For third-generation dual source CT (DSCT), tube voltage combinations of 150Sn and 70, 80, 90, 100 kVp were analysed. For dual layer CT (DLCT), 120 and 140 kVp were used. Scans were repeated three times. Median normalized values and interquartile ranges (IQRs) were calculated for all kVp settings and phantom sizes. Correlation between measured and known iodine concentrations was excellent for both systems (R = 0.999-1.000, p < 0.0001). For DSCT, median measurement errors ranged from -0.5% (IQR -2.0, 2.0%) at 150Sn/70 kVp and -2.3% (IQR -4.0, -0.1%) at 150Sn/80 kVp to -4.0% (IQR -6.0, -2.8%) at 150Sn/90 kVp. For DLCT, median measurement errors ranged from -3.3% (IQR -4.9, -1.5%) at 140 kVp to -4.6% (IQR -6.0, -3.6%) at 120 kVp. Larger phantom sizes increased variability of iodine measurements (p < 0.05). Iodine concentration can be accurately quantified with state-of-the-art DECT systems from two vendors. The lowest absolute errors were found for DSCT using the 150Sn/70 kVp or 150Sn/80 kVp combinations, which was slightly more accurate than 140 kVp in DLCT. • High-end CT scanners allow accurate iodine quantification using different DECT techniques. • Lowest measurement error was found in scans with largest photon energy separation. • Dual-source CT quantified iodine slightly more accurately than dual layer CT.

  3. Diagnostic accuracy of low-dose CT compared with abdominal radiography in non-traumatic acute abdominal pain: prospective study and systematic review.

    Science.gov (United States)

    Alshamari, Muhammed; Norrman, Eva; Geijer, Mats; Jansson, Kjell; Geijer, Håkan

    2016-06-01

    Abdominal radiography is frequently used in acute abdominal non-traumatic pain despite the availability of more advanced diagnostic modalities. This study evaluates the diagnostic accuracy of low-dose CT compared with abdominal radiography, at similar radiation dose levels. Fifty-eight patients were imaged with both methods and were reviewed independently by three radiologists. The reference standard was obtained from the diagnosis in medical records. Sensitivity and specificity were calculated. A systematic review was performed after a literature search, finding a total of six relevant studies including the present. Overall sensitivity with 95 % CI for CT was 75 % (66-83 %) and 46 % (37-56 %) for radiography. Specificity was 87 % (77-94 %) for both methods. In the systematic review the overall sensitivity for CT varied between 75 and 96 % with specificity from 83 to 95 % while the overall sensitivity for abdominal radiography varied between 30 and 77 % with specificity 75 to 88 %. Based on the current study and available evidence, low-dose CT has higher diagnostic accuracy than abdominal radiography and it should, where logistically possible, replace abdominal radiography in the workup of adult patients with acute non-traumatic abdominal pain. • Low-dose CT has a higher diagnostic accuracy than radiography. • A systematic review shows that CT has better diagnostic accuracy than radiography. • Radiography has no place in the workup of acute non-traumatic abdominal pain.

  4. Accuracy and efficiency of radiologists reporting CT studies from a multiscreen electronic workstation

    International Nuclear Information System (INIS)

    Foley, W.D.; Lawson, T.L.; Erickson, S.J.

    1988-01-01

    This paper describes a limited picture archiving communication system linking, a via fiber optic cable, two computed tomographic (CT) scanners, and a double-screen (1,000-line monitors) display station with a 600-megabyte image memory evaluated for accuracy and efficiency in the generation of a radiology report. Twenty preselected CT studies (an average of 30 images per study, ten with prior examination for comparison) in cases with pathologically or clinically proved diagnoses were read by five radiologists, in random order, from both the electronic display console and conventional multiformat images. All cases were reread after 1-month interval in a different random order viewing sequence and the opposite viewing modality

  5. Improving the accuracy of CT dimensional metrology by a novel beam hardening correction method

    International Nuclear Information System (INIS)

    Zhang, Xiang; Li, Lei; Zhang, Feng; Xi, Xiaoqi; Deng, Lin; Yan, Bin

    2015-01-01

    Its powerful nondestructive characteristics are attracting more and more research into the study of computed tomography (CT) for dimensional metrology, which offers a practical alternative to the common measurement methods. However, the inaccuracy and uncertainty severely limit the further utilization of CT for dimensional metrology due to many factors, among which the beam hardening (BH) effect plays a vital role. This paper mainly focuses on eliminating the influence of the BH effect in the accuracy of CT dimensional metrology. To correct the BH effect, a novel exponential correction model is proposed. The parameters of the model are determined by minimizing the gray entropy of the reconstructed volume. In order to maintain the consistency and contrast of the corrected volume, a punishment term is added to the cost function, enabling more accurate measurement results to be obtained by the simple global threshold method. The proposed method is efficient, and especially suited to the case where there is a large difference in gray value between material and background. Different spheres with known diameters are used to verify the accuracy of dimensional measurement. Both simulation and real experimental results demonstrate the improvement in measurement precision. Moreover, a more complex workpiece is also tested to show that the proposed method is of general feasibility. (paper)

  6. Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of pediatric abdominal CT

    Energy Technology Data Exchange (ETDEWEB)

    Bae, Sohi; Kim, Myung-Joon; Lee, Mi-Jung [Yonsei University College of Medicine, Department of Radiology and Research Institute of Radiological Science, Severance Children' s Hospital, Seoul (Korea, Republic of); Yoon, Choon-Sik [Yonsei University College of Medicine, Department of Radiology, Gangnam Severance Hospital, Seoul (Korea, Republic of); Kim, Dong Wook; Hong, Jung Hwa [Yonsei University College of Medicine, Biostatistics Collaboration Unit, Seoul (Korea, Republic of)

    2014-12-15

    Since children are more radio-sensitive than adults, there is a need to minimize radiation exposure during CT exams. To evaluate the effects of adaptive statistical iterative reconstruction (ASIR) on radiation dose reduction, image quality and diagnostic accuracy in pediatric abdominal CT. We retrospectively reviewed the abdominal CT examinations of 41 children (24 boys and 17 girls; mean age: 10 years) with a low-dose radiation protocol and reconstructed with ASIR (the ASIR group). We also reviewed routine-dose abdominal CT examinations of 41 age- and sex-matched controls reconstructed with filtered-back projection (control group). Image quality was assessed objectively as noise measured in the liver, spleen and aorta, as well as subjectively by three pediatric radiologists for diagnostic acceptability using a four-point scale. Radiation dose and objective image qualities of each group were compared with the paired t-test. Diagnostic accuracy was evaluated by reviewing follow-up imaging studies and medical records in 2012 and 2013. There was 46.3% dose reduction of size-specific dose estimates in ASIR group (from 13.4 to 7.2 mGy) compared with the control group. Objective noise was higher in the liver, spleen and aorta of the ASIR group (P < 0.001). However, the subjective image quality was average or superior in 84-100% of studies. Only one image was subjectively rated as unacceptable by one reviewer. There was only one case with interpretational error in the control group and none in the ASIR group. Use of the ASIR technique resulted in greater than a 45% reduction in radiation dose without impairing subjective image quality or diagnostic accuracy in pediatric abdominal CT, despite increased objective image noise. (orig.)

  7. Effects of adaptive statistical iterative reconstruction on radiation dose reduction and diagnostic accuracy of pediatric abdominal CT

    International Nuclear Information System (INIS)

    Bae, Sohi; Kim, Myung-Joon; Lee, Mi-Jung; Yoon, Choon-Sik; Kim, Dong Wook; Hong, Jung Hwa

    2014-01-01

    Since children are more radio-sensitive than adults, there is a need to minimize radiation exposure during CT exams. To evaluate the effects of adaptive statistical iterative reconstruction (ASIR) on radiation dose reduction, image quality and diagnostic accuracy in pediatric abdominal CT. We retrospectively reviewed the abdominal CT examinations of 41 children (24 boys and 17 girls; mean age: 10 years) with a low-dose radiation protocol and reconstructed with ASIR (the ASIR group). We also reviewed routine-dose abdominal CT examinations of 41 age- and sex-matched controls reconstructed with filtered-back projection (control group). Image quality was assessed objectively as noise measured in the liver, spleen and aorta, as well as subjectively by three pediatric radiologists for diagnostic acceptability using a four-point scale. Radiation dose and objective image qualities of each group were compared with the paired t-test. Diagnostic accuracy was evaluated by reviewing follow-up imaging studies and medical records in 2012 and 2013. There was 46.3% dose reduction of size-specific dose estimates in ASIR group (from 13.4 to 7.2 mGy) compared with the control group. Objective noise was higher in the liver, spleen and aorta of the ASIR group (P < 0.001). However, the subjective image quality was average or superior in 84-100% of studies. Only one image was subjectively rated as unacceptable by one reviewer. There was only one case with interpretational error in the control group and none in the ASIR group. Use of the ASIR technique resulted in greater than a 45% reduction in radiation dose without impairing subjective image quality or diagnostic accuracy in pediatric abdominal CT, despite increased objective image noise. (orig.)

  8. Improvement of analysis precision upon the atomic number and electron density measurement by the dual x-ray CT

    Science.gov (United States)

    Imura, Yukino; Morii, Hisashi; Koike, Akifumi; Okunoyama, Takaharu; Neo, Yoichiro; Mimura, Hidenori; Aoki, Toru

    2010-08-01

    To identify the factor impairing the material identification parameters, which is provided by the dual-energy X-ray computed tomography method using a conventional X-ray tube and a CdTe detector, linear attenuation coefficient was measured by the radioactivity of radio isotopes and compared with theoretical figure. In our study, the atomic number and the electron density is calculated from the linear attenuation coefficient obtained in CT measurement by 64-channel CdTe line detector. To estimate accuracy of CdTe line sensor, it is needed to obtain the linear attenuation coefficient accurately. Using a single detector, the linear attenuation coefficient is verified for accuracy. The energy resolution of CdTe detectors and the method of reconstruction are discussed.

  9. Combined FDG PET/CT imaging for restaging of colorectal cancer patients: impact of image fusion on staging accuracy

    International Nuclear Information System (INIS)

    Strunk, H.; Jaeger, U.; Flacke, S.; Hortling, N.; Bucerius, J.; Joe, A.; Reinhardt, M.; Palmedo, H.

    2005-01-01

    Purpose: To evaluate the diagnostic impact of positron emission tomography (PET) with fluorine-18-labeled deoxy-D-glucose (FDG) combined with non-contrast computed tomography (CT) as PET-CT modality in restaging colorectal cancer patients. Material and methods: In this retrospective study, 29 consecutive patients with histologically proven colorectal cancer (17 female, 12 male, aged 51-76 years) underwent whole body scans in one session on a dual modality PET-CT system (Siemens Biograph) 90 min. after i.v. administration of 370 MBq 18 F-FDG. The CT imaging was performed with 40 mAs, 130 kV, slice-thickness 5 mm and without i.v. contrast administration. PET and CT images were reconstructed with a slice-thickness of 5 mm in coronal, sagittal and transverse planes. During a first step of analysis, PET and CT images were scored blinded and independently by a group of two nuclear medicine physicians and a group of two radiologists, respectively. For this purpose, a five-point-scale was used. The second step of data-analysis consisted of a consensus reading by both groups. During the consensus reading, first a virtual (meaning mental) fusion of PET and CT images and afterwards the 'real' fusion (meaning coregistered) PET-CT images were also scored with the same scale. The imaging results were compared with histopathology findings and the course of disease during further follow-up. Results: The total number of malignant lesions detected with the combined PET/CT were 86. For FDG-PET alone it was n=68, and for CT alone n=65. Comparing PET-CT and PET, concordance was found in 81 of 104 lesions. Discrepancies predominantly occurred in the lung, where PET alone often showed true positive results in lymph nodes and soft tissue masses, where CT often was false negative. Comparing mental fusion and 'real' co-registered images, concordance was found in 94 of 104 lesions. In 13 lesions or, respectively, in 7 of 29 patients, a relevant information was gathered using fused images

  10. TU-F-18A-09: CT Number Stability Across Patient Sizes Using Virtual-Monoenergetic Dual-Energy CT

    Energy Technology Data Exchange (ETDEWEB)

    Michalak, G; Grimes, J; Fletcher, J; McCollough, C [Mayo Clinic, Rochester, MN (United States); Halaweish, A [Siemens Healthcare, Rochester, MN (United States)

    2014-06-15

    Purpose: Virtual-monoenergetic imaging uses dual-energy CT data to synthesize images corresponding to a single photon energy, thereby reducing beam-hardening artifacts. This work evaluated the ability of a commercial virtual-monoenergetic algorithm to achieve stable CT numbers across patient sizes. Methods: Test objects containing a range of iodine and calcium hydroxyapatite concentrations were placed inside 8 torso-shaped water phantoms, ranging in lateral width from 15 to 50 cm, and scanned on a dual-source CT system (Siemens Somatom Force). Single-energy scans were acquired from 70-150 kV in 10 kV increments; dual-energy scans were acquired using 4 energy pairs (low energy: 70, 80, 90, and 100 kV; high energy: 150 kV + 0.6 mm Sn). CTDIvol was matched for all single- and dual-energy scans for a given phantom size. All scans used 128×0.6 mm collimation and were reconstructed with 1-mm thickness at 0.8-mm increment and a medium smooth body kernel. Monoenergetic images were generated using commercial software (syngo Via Dual Energy, VA30). Iodine contrast was calculated as the difference in mean iodine and water CT numbers from respective regions-of-interest in 10 consecutive images. Results: CT numbers remained stable as phantom width varied from 15 to 50 cm for all dual-energy data sets (except for at 50 cm using 70/150Sn due to photon starvation effects). Relative to the 15 cm phantom, iodine contrast was within 5.2% of the 70 keV value for phantom sizes up to 45 cm. At 90/150Sn, photon starvation did not occur at 50 cm, and iodine contrast in the 50-cm phantom was within 1.4% of the 15-cm phantom. Conclusion: Monoenergetic imaging, as implemented in the evaluated commercial system, eliminated the variation in CT numbers due to patient size, and may provide more accurate data for quantitative tasks, including radiation therapy treatment planning. Siemens Healthcare.

  11. Para-aortic lymphadenectomy in advanced stage cervical cancer, a protocol for comparing safety, feasibility and diagnostic accuracy of surgical staging versus PET-CT; PALDISC trial.

    Science.gov (United States)

    Tax, Casper; Abbink, Karin; Rovers, Maroeska M; Bekkers, Ruud L M; Zusterzeel, Petra L M

    2018-01-01

    Currently, a PET-CT is used to assess the need for extended field radiotherapy of para-aortic lymph nodes (PALN) in International Federation of Gynaecology and Obstetrics (FIGO) stage IB2, IIA2-IVA (locally advanced stage) cervical cancer. A small study established a sensitivity and specificity estimate for PALN metastases of 50% (95% CI; 7-93%) and 83% (95% CI; 52-98%), respectively. Surgical staging of PALN may lead to a higher diagnostic accuracy. However, surgical staging of para-aortic lymph nodes in locally advanced stage cervical cancer is not common practice. Therefore, a phase 2 randomised controlled trial is needed to assess its safety and feasibility. In addition to standard imaging (MRI or CT scan) with PET-CT, 30 adult women with FIGO stage IB2, IIA2-IVA cervical cancer will be randomised to receive either surgical staging or usual PET-CT staging. Administering extended field radiotherapy will be based on lymphadenectomy results for the intervention group and on the PET-CT results for the control group. Follow-up visits at 0, 3, 6, 9 and 12 months will assess health-related quality of life and progression-free survival.Primary safety and feasibility outcomes of surgical staging will be assessed by calculating means with 95% confidence intervals for duration of surgery, number of complications, blood loss, nodal yield after para-aortic lymphadenectomy and treatment delay due to surgical staging. Secondary patient-centred outcomes on quality of life and first year survival will be documented and compared between the two groups. Estimates of sensitivity, specificity and negative and positive predictive values of MRI, PET-CT and surgical staging will be presented with 95% CI.. All analysis will be performed according to the intention to treat principle. This study will assess safety and feasibility, expressed as the number and severity of complications, effect on quality of life and the treatment delay due to surgically staging para-aortic lymph nodes in

  12. High Dose MicroCT Does Not Contribute Toward Improved MicroPET/CT Image Quantitative Accuracy and Can Limit Longitudinal Scanning of Small Animals

    Directory of Open Access Journals (Sweden)

    Wendy A. McDougald

    2017-10-01

    Full Text Available Obtaining accurate quantitative measurements in preclinical Positron Emission Tomography/Computed Tomography (PET/CT imaging is of paramount importance in biomedical research and helps supporting efficient translation of preclinical results to the clinic. The purpose of this study was two-fold: (1 to investigate the effects of different CT acquisition protocols on PET/CT image quality and data quantification; and (2 to evaluate the absorbed dose associated with varying CT parameters.Methods: An air/water quality control CT phantom, tissue equivalent material phantom, an in-house 3D printed phantom and an image quality PET/CT phantom were imaged using a Mediso nanoPET/CT scanner. Collected data was analyzed using PMOD software, VivoQuant software and National Electric Manufactures Association (NEMA software implemented by Mediso. Measured Hounsfield Unit (HU in collected CT images were compared to the known HU values and image noise was quantified. PET recovery coefficients (RC, uniformity and quantitative bias were also measured.Results: Only less than 2 and 1% of CT acquisition protocols yielded water HU values < −80 and air HU values < −840, respectively. Four out of 11 CT protocols resulted in more than 100 mGy absorbed dose. Different CT protocols did not impact PET uniformity and RC, and resulted in <4% overall bias relative to expected radioactive concentration.Conclusion: Preclinical CT protocols with increased exposure times can result in high absorbed doses to the small animals. These should be avoided, as they do not contributed toward improved microPET/CT image quantitative accuracy and could limit longitudinal scanning of small animals.

  13. A Simplified Whole-Organ CT Perfusion Technique with Biphasic Acquisition: Preliminary Investigation of Accuracy and Protocol Feasibility in Kidneys.

    Science.gov (United States)

    Yuan, XiaoDong; Zhang, Jing; Quan, ChangBin; Tian, Yuan; Li, Hong; Ao, GuoKun

    2016-04-01

    To determine the feasibility and accuracy of a protocol for calculating whole-organ renal perfusion (renal blood flow [RBF]) and regional perfusion on the basis of biphasic computed tomography (CT), with concurrent dynamic contrast material-enhanced (DCE) CT perfusion serving as the reference standard. This prospective study was approved by the institutional review board, and written informed consent was obtained from all patients. Biphasic CT of the kidneys, including precontrast and arterial phase imaging, was integrated with a first-pass dynamic volume CT protocol and performed and analyzed in 23 patients suspected of having renal artery stenosis. The perfusion value derived from biphasic CT was calculated as CT number enhancement divided by the area under the arterial input function and compared with the DCE CT perfusion data by using the paired t test, correlation analysis, and Bland-Altman plots. Correlation analysis was made between the RBF and the extent of renal artery stenosis. All postprocessing was independently performed by two observers and then averaged as the final result. Mean ± standard deviation biphasic and DCE CT perfusion data for RBF were 425.62 mL/min ± 124.74 and 419.81 mL/min ± 121.13, respectively (P = .53), and for regional perfusion they were 271.15 mL/min per 100 mL ± 82.21 and 266.33 mL/min per 100 mL ± 74.40, respectively (P = .31). Good correlation and agreement were shown between biphasic and DCE CT perfusion for RBF (r = 0.93; ±10% variation from mean perfusion data [P < .001]) and for regional perfusion (r = 0.90; ±13% variation from mean perfusion data [P < .001]). The extent of renal artery stenosis was negatively correlated with RBF with biphasic CT perfusion (r = -0.81, P = .012). Biphasic CT perfusion is clinically feasible and provides perfusion data comparable to DCE CT perfusion data at both global and regional levels in the kidney. Online supplemental material is available for this article.

  14. Low density lesion in solid mass on CT: Pathologic change and housfield number

    International Nuclear Information System (INIS)

    Han, Tae Il; Lim, Joo Won; Ryu, Kyung Nam; Ko, Young Tae; Song, Mi Jin; Lee, Dong Ho; Lee, Ju Hie

    1994-01-01

    We retrospectively reviewed the pathologic changes and housfield unit of the low density lesion in solid mass on CT. Pathologically proved solid mass was evaluated in regard to the shape and margin of the low density in the mass on the CT scans of 23 patient. The CT number of the low density lesion was correlated with the pathologic changes. Pathologic changes of the low density lesions were; necrosis (n=17), hemorrhage (n=13), cyst (n=4), myxoid degeneration (n=2), hyaline degeneration (n=1), fibrosis (n=1), and mixed cellularity (n=1). In 14 cases, more than 2 pathologic changes were seen. In 11 cases, necrosis was associated with hemorrhage. The CT number ranged from 11.5 to 44.9 Housfield unit(HU) (mean, 25.2 HU). The average CT number was 26.9 HU in hemorrhage and necrosis, 17.2 HU in cystic change, 20.9 HU in myxoid degeneration, 35.7 HU in hyaline de generation, 22.3 HU in fibrosis, and 21.4 HU in mixed cellularity. The hemorrhage and necrosis in 17 cases showed irregular margin, amorphous shape, and showed centrifugal distribution. The cystic change in 4 cases showed well defined margin, round shape, and peripheral location in solid mass. The low density lesions in solid mass on CT represented variable pathologic changes; necrosis, hemorrhage, cyst, myxoid degeneration, hyaline degeneration, fibrosis, and mixed cellularity. Pathologic changes would not be differentiated on the basis of CT number

  15. Added value of cost-utility analysis in simple diagnostic studies of accuracy: (18)F-fluoromethylcholine PET/CT in prostate cancer staging.

    Science.gov (United States)

    Gerke, Oke; Poulsen, Mads H; Høilund-Carlsen, Poul Flemming

    2015-01-01

    Diagnostic studies of accuracy targeting sensitivity and specificity are commonly done in a paired design in which all modalities are applied in each patient, whereas cost-effectiveness and cost-utility analyses are usually assessed either directly alongside to or indirectly by means of stochastic modeling based on larger randomized controlled trials (RCTs). However the conduct of RCTs is hampered in an environment such as ours, in which technology is rapidly evolving. As such, there is a relatively limited number of RCTs. Therefore, we investigated as to which extent paired diagnostic studies of accuracy can be also used to shed light on economic implications when considering a new diagnostic test. We propose a simple decision tree model-based cost-utility analysis of a diagnostic test when compared to the current standard procedure and exemplify this approach with published data from lymph node staging of prostate cancer. Average procedure costs were taken from the Danish Diagnosis Related Groups Tariff in 2013 and life expectancy was estimated for an ideal 60 year old patient based on prostate cancer stage and prostatectomy or radiation and chemotherapy. Quality-adjusted life-years (QALYs) were deduced from the literature, and an incremental cost-effectiveness ratio (ICER) was used to compare lymph node dissection with respective histopathological examination (reference standard) and (18)F-fluoromethylcholine positron emission tomography/computed tomography (FCH-PET/CT). Lower bounds of sensitivity and specificity of FCH-PET/CT were established at which the replacement of the reference standard by FCH-PET/CT comes with a trade-off between worse effectiveness and lower costs. Compared to the reference standard in a diagnostic accuracy study, any imperfections in accuracy of a diagnostic test imply that replacing the reference standard generates a loss in effectiveness and utility. We conclude that diagnostic studies of accuracy can be put to a more extensive use

  16. Identification of Nasal Bone Fractures on Conventional Radiography and Facial CT: Comparison of the Diagnostic Accuracy in Different Imaging Modalities and Analysis of Interobserver Reliability

    International Nuclear Information System (INIS)

    Baek, Hye Jin; Kim, Dong Wook; Ryu, Ji Hwa; Lee, Yoo Jin

    2013-01-01

    There has been no study to compare the diagnostic accuracy of an experienced radiologist with a trainee in nasal bone fracture. To compare the diagnostic accuracy between conventional radiography and computed tomography (CT) for the identification of nasal bone fractures and to evaluate the interobserver reliability between a staff radiologist and a trainee. A total of 108 patients who underwent conventional radiography and CT after acute nasal trauma were included in this retrospective study. Two readers, a staff radiologist and a second-year resident, independently assessed the results of the imaging studies. Of the 108 patients, the presence of a nasal bone fracture was confirmed in 88 (81.5%) patients. The number of non-depressed fractures was higher than the number of depressed fractures. In nine (10.2%) patients, nasal bone fractures were only identified on conventional radiography, including three depressed and six non-depressed fractures. CT was more accurate as compared to conventional radiography for the identification of nasal bone fractures as determined by both readers (P <0.05), all diagnostic indices of an experienced radiologist were similar to or higher than those of a trainee, and κ statistics showed moderate agreement between the two diagnostic tools for both readers. There was no statistical difference in the assessment of interobserver reliability for both imaging modalities in the identification of nasal bone fractures. For the identification of nasal bone fractures, CT was significantly superior to conventional radiography. Although a staff radiologist showed better values in the identification of nasal bone fracture and differentiation between depressed and non-depressed fractures than a trainee, there was no statistically significant difference in the interpretation of conventional radiography and CT between a radiologist and a trainee

  17. Study of Image Quality From CT Scanner Multi-Detector by using Americans College of Radiology (ACR) Phantom

    Science.gov (United States)

    Mulyadin; Dewang, Syamsir; Abdullah, Bualkar; Tahir, Dahlang

    2018-03-01

    In this study, the image quality of CT scan using phantom American College of Radiology (ACR) was determined. Scanning multidetector CT is used to know the image quality parameters by using a solid phantom containing four modules and primarily from materials that are equivalent to water. Each module is 4 cm in diameter and 20 cm in diameter. There is white alignment marks painted white to reflect the alignment laser and there are also “HEAD”, “FOOT”, and “TOP” marks on the phantom to help align. This test obtains CT images of each module according to the routine inspection protocol of the head. Acceptance of image quality obtained for determination: CT Number Accuracy (CTN), CT Number Uniformity and Noise, Linearity CT Number, Slice Technique, Low Contrast Resolution and High Contrast Resolution represent image quality parameters. In testing CT Number Accuracy (CTN), CT Uniform number and Noise are in the range of tolerable values allowed. In the test, Linearity CT Number obtained correlation value above 0.99 is the relationship between electron density and CT Number. In a low contrast resolution test, the smallest contrast groups are visible. In contrast, the high resolution is seen up to 7 lp/cm. The quality of GE CT Scan is very high, as all the image quality tests obtained are within the tolerance brackets of values permitted by the Nuclear Power Control Agency (BAPETEN). Image quality test is a way to get very important information about the accuracy of snoring result by using phantom ACR.

  18. An approach for quantitative image quality analysis for CT

    Science.gov (United States)

    Rahimi, Amir; Cochran, Joe; Mooney, Doug; Regensburger, Joe

    2016-03-01

    An objective and standardized approach to assess image quality of Compute Tomography (CT) systems is required in a wide variety of imaging processes to identify CT systems appropriate for a given application. We present an overview of the framework we have developed to help standardize and to objectively assess CT image quality for different models of CT scanners used for security applications. Within this framework, we have developed methods to quantitatively measure metrics that should correlate with feature identification, detection accuracy and precision, and image registration capabilities of CT machines and to identify strengths and weaknesses in different CT imaging technologies in transportation security. To that end we have designed, developed and constructed phantoms that allow for systematic and repeatable measurements of roughly 88 image quality metrics, representing modulation transfer function, noise equivalent quanta, noise power spectra, slice sensitivity profiles, streak artifacts, CT number uniformity, CT number consistency, object length accuracy, CT number path length consistency, and object registration. Furthermore, we have developed a sophisticated MATLAB based image analysis tool kit to analyze CT generated images of phantoms and report these metrics in a format that is standardized across the considered models of CT scanners, allowing for comparative image quality analysis within a CT model or between different CT models. In addition, we have developed a modified sparse principal component analysis (SPCA) method to generate a modified set of PCA components as compared to the standard principal component analysis (PCA) with sparse loadings in conjunction with Hotelling T2 statistical analysis method to compare, qualify, and detect faults in the tested systems.

  19. {sup 68}Ga-DOTATATE PET/CT in recurrent medullary thyroid carcinoma: a lesion-by-lesion comparison with {sup 111}In-octreotide SPECT/CT and conventional imaging

    Energy Technology Data Exchange (ETDEWEB)

    Yamaga, Lilian Yuri Itaya; Cunha, Marcelo L.; Campos Neto, Guilherme C.; Garcia, Marcio R.T.; Wagner, Jairo; Funari, Marcelo B.G. [Hospital Israelita Albert Einstein, Imaging Department, Sao Paulo (Brazil); Yang, Ji H.; Camacho, Cleber P. [Universidade Federal de Sao Paulo, Multiple Neoplasia Outpatiet Clinic, Sao Paulo (Brazil)

    2017-09-15

    The aim of this study was to prospectively compare the detection rate of {sup 68}Ga-DOTATATE PET-CT with {sup 111}In-octreotide SPECT-CT and conventional imaging (CI) in medullary thyroid carcinoma (MTC) patients with increased calcitonin (Ctn) levels but negative CI after thyroidectomy. Fifteen patients with raised Ctn levels and/or CI evidence of recurrence underwent {sup 68}Ga-DOTATATE PET-CT, {sup 111}In-octreotide SPECT-CT and CI. Histopathology, CI and biochemical/clinical/imaging follow-up were used as the reference standard. PET/CT, SPECT/CT and CI were compared in a lesion-based and organ-based analysis. PET/CT evidenced recurrence in 14 of 15 patients. There were 13 true positive (TP), 1 true negative (TN), 1 false positive (FP) and no false negative (FN) cases, resulting in a sensitivity and accuracy of 100% and 93%. SPECT/CT was positive in 6 of 15 cases. There were 6 TP, 2 TN, 7 FN and no FP cases, resulting in a sensitivity of 46% and accuracy of 53%. CI procedures detected tumor lesions in 14 of 15 patients. There were 13 TP, 1TN, 1 FP and no FN cases with a sensitivity of 100% and accuracy of 93%. A significantly higher number of lesions was detected by PET/CT (112 lesions, p = 0.005) and CI (109 lesions, p = 0.005) in comparison to SPECT/CT (16 lesions). There was no significant difference between PET/CT and CI for the total number of detected lesions (p = 0.734). PET/CT detected more lesions than SPECT/CT regardless of the organ. PET/CT detected more bone lesions but missed some neck nodal metastases evidenced by CI. The number of lesions per region demonstrated by PET/CT and CI were similar in the other sites. {sup 68}Ga-DOTATATE PET/CT is superior to {sup 111}In-octreotide SPECT/CT for the detection of recurrent MTC demonstrating a significantly higher number of lesions. {sup 68}Ga-DOTATATE PET/CT showed a superior detection rate compared to CI in demonstrating bone metastases. (orig.)

  20. Comparison of CT number calibration techniques for CBCT-based dose calculation

    Energy Technology Data Exchange (ETDEWEB)

    Dunlop, Alex [The Royal Marsden NHS Foundation Trust, Joint Department of Physics, Institute of Cancer Research, London (United Kingdom); The Royal Marsden Hospital, Sutton, Surrey, Downs Road (United Kingdom); McQuaid, Dualta; Nill, Simeon; Hansen, Vibeke N.; Oelfke, Uwe [The Royal Marsden NHS Foundation Trust, Joint Department of Physics, Institute of Cancer Research, London (United Kingdom); Murray, Julia; Bhide, Shreerang; Harrington, Kevin [The Royal Marsden Hospital, Sutton, Surrey, Downs Road (United Kingdom); The Institute of Cancer Research, London (United Kingdom); Poludniowski, Gavin [Karolinska University Hospital, Department of Medical Physics, Stockholm (Sweden); Nutting, Christopher [The Institute of Cancer Research, London (United Kingdom); Newbold, Kate [The Royal Marsden Hospital, Sutton, Surrey, Downs Road (United Kingdom)

    2015-12-15

    The aim of this work was to compare and validate various computed tomography (CT) number calibration techniques with respect to cone beam CT (CBCT) dose calculation accuracy. CBCT dose calculation accuracy was assessed for pelvic, lung, and head and neck (H and N) treatment sites for two approaches: (1) physics-based scatter correction methods (CBCT{sub r}); (2) density override approaches including assigning water density to the entire CBCT (W), assignment of either water or bone density (WB), and assignment of either water or lung density (WL). Methods for CBCT density assignment within a commercially available treatment planning system (RS{sub auto}), where CBCT voxels are binned into six density levels, were assessed and validated. Dose-difference maps and dose-volume statistics were used to compare the CBCT dose distributions with the ground truth of a planning CT acquired the same day as the CBCT. For pelvic cases, all CTN calibration methods resulted in average dose-volume deviations below 1.5 %. RS{sub auto} provided larger than average errors for pelvic treatments for patients with large amounts of adipose tissue. For H and N cases, all CTN calibration methods resulted in average dose-volume differences below 1.0 % with CBCT{sub r} (0.5 %) and RS{sub auto} (0.6 %) performing best. For lung cases, WL and RS{sub auto} methods generated dose distributions most similar to the ground truth. The RS{sub auto} density override approach is an attractive option for CTN adjustments for a variety of anatomical sites. RS{sub auto} methods were validated, resulting in dose calculations that were consistent with those calculated on diagnostic-quality CT images, for CBCT images acquired of the lung, for patients receiving pelvic RT in cases without excess adipose tissue, and for H and N cases. (orig.) [German] Ziel dieser Arbeit ist der Vergleich und die Validierung mehrerer CT-Kalibrierungsmethoden zur Dosisberechnung auf der Grundlage von Kegelstrahlcomputertomographie

  1. CTC-ask: a new algorithm for conversion of CT numbers to tissue parameters for Monte Carlo dose calculations applying DICOM RS knowledge

    International Nuclear Information System (INIS)

    Ottosson, Rickard O; Behrens, Claus F

    2011-01-01

    One of the building blocks in Monte Carlo (MC) treatment planning is to convert patient CT data to MC compatible phantoms, consisting of density and media matrices. The resulting dose distribution is highly influenced by the accuracy of the conversion. Two major contributing factors are precise conversion of CT number to density and proper differentiation between air and lung. Existing tools do not address this issue specifically. Moreover, their density conversion may depend on the number of media used. Differentiation between air and lung is an important task in MC treatment planning and misassignment may lead to local dose errors on the order of 10%. A novel algorithm, CTC-ask, is presented in this study. It enables locally confined constraints for the media assignment and is independent of the number of media used for the conversion of CT number to density. MC compatible phantoms were generated for two clinical cases using a CT-conversion scheme implemented in both CTC-ask and the DICOM-RT toolbox. Full MC dose calculation was subsequently conducted and the resulting dose distributions were compared. The DICOM-RT toolbox inaccurately assigned lung in 9.9% and 12.2% of the voxels located outside of the lungs for the two cases studied, respectively. This was completely avoided by CTC-ask. CTC-ask is able to reduce anatomically irrational media assignment. The CTC-ask source code can be made available upon request to the authors. (note)

  2. Accuracy of computed tomographic intravenous cholangiography (CT-IVC) with iotroxate in the detection of choledocholithiasis

    International Nuclear Information System (INIS)

    Gibson, Robert N.; Vincent, Janette M.; Speer, Tony; Noack, Keith; Collier, Neil A.

    2005-01-01

    To determine the accuracy of computed tomographic intravenous cholangiography (CT-IVC) for detection of choledocholithiasis. Sixty-five patients undergoing endoscopic retrograde cholangiography (ERC) underwent CT-IVC prior to ERC, using a single detector helical CT following intravenous infusion of 100 ml iotroxate. Patients with bilirubin levels >3 times normal were excluded. ERC was indeterminate in three patients (4.7%) and CT-IVC in four (6.3%). Twenty-three patients had ductal calculi at ERC, and CT-IVC was positive in 22, with two false positives and one false negative: sensitivity 95.6%, specificity 94.3%. Stones were multiple in nine and solitary in 14. Of the 14 solitary stones, ten were ≤5 mm and eight were ≤4 mm. The bilirubin level in positive cases was within normal in 20. Maximum intensity projection (MIP) reformats showed stones in only 27% of cases and surface rendered (SR) reformats in none. CT-IVC is highly accurate for detection of ductal calculi, including single small calculi, with a normal or near normal serum bilirubin. Axial images should be used for interpretation rather than MIP or SR reformats. (orig.)

  3. Learning linear spatial-numeric associations improves accuracy of memory for numbers

    Directory of Open Access Journals (Sweden)

    Clarissa Ann Thompson

    2016-01-01

    Full Text Available Memory for numbers improves with age and experience. One potential source of improvement is a logarithmic-to-linear shift in children’s representations of magnitude. To test this, Kindergartners and second graders estimated the location of numbers on number lines and recalled numbers presented in vignettes (Study 1. Accuracy at number-line estimation predicted memory accuracy on a numerical recall task after controlling for the effect of age and ability to approximately order magnitudes (mapper status. To test more directly whether linear numeric magnitude representations caused improvements in memory, half of children were given feedback on their number-line estimates (Study 2. As expected, learning linear representations was again linked to memory for numerical information even after controlling for age and mapper status. These results suggest that linear representations of numerical magnitude may be a causal factor in development of numeric recall accuracy.

  4. Reduction of Cone-Beam CT scan time without compromising the accuracy of the image registration in IGRT

    DEFF Research Database (Denmark)

    Westberg, Jonas; Jensen, Henrik R; Bertelsen, Anders

    2010-01-01

    In modern radiotherapy accelerators are equipped with 3D cone-beam CT (CBCT) which is used to verify patient position before treatment. The verification is based on an image registration between the CBCT acquired just before treatment and the CT scan made for the treatment planning. The purpose...... of this study is to minimise the scan time of the CBCT without compromising the accuracy of the image registration in IGRT....

  5. Application of CT-PSF-based computer-simulated lung nodules for evaluating the accuracy of computer-aided volumetry.

    Science.gov (United States)

    Funaki, Ayumu; Ohkubo, Masaki; Wada, Shinichi; Murao, Kohei; Matsumoto, Toru; Niizuma, Shinji

    2012-07-01

    With the wide dissemination of computed tomography (CT) screening for lung cancer, measuring the nodule volume accurately with computer-aided volumetry software is increasingly important. Many studies for determining the accuracy of volumetry software have been performed using a phantom with artificial nodules. These phantom studies are limited, however, in their ability to reproduce the nodules both accurately and in the variety of sizes and densities required. Therefore, we propose a new approach of using computer-simulated nodules based on the point spread function measured in a CT system. The validity of the proposed method was confirmed by the excellent agreement obtained between computer-simulated nodules and phantom nodules regarding the volume measurements. A practical clinical evaluation of the accuracy of volumetry software was achieved by adding simulated nodules onto clinical lung images, including noise and artifacts. The tested volumetry software was revealed to be accurate within an error of 20 % for nodules >5 mm and with the difference between nodule density and background (lung) (CT value) being 400-600 HU. Such a detailed analysis can provide clinically useful information on the use of volumetry software in CT screening for lung cancer. We concluded that the proposed method is effective for evaluating the performance of computer-aided volumetry software.

  6. Prediction of nodal involvement in primary rectal carcinoma without invasion to pelvic structures: accuracy of preoperative CT, MR, and DWIBS assessments relative to histopathologic findings.

    Directory of Open Access Journals (Sweden)

    Jun Zhou

    Full Text Available OBJECTIVE: To investigate the accuracy of preoperative computed tomography (CT, magnetic resonance (MR imaging and diffusion-weighted imaging with background body signal suppression (DWIBS in the prediction of nodal involvement in primary rectal carcinoma patients in the absence of tumor invasion into pelvic structures. METHODS AND MATERIALS: Fifty-two subjects with primary rectal cancer were preoperatively assessed by CT and MRI at 1.5 T with a phased-array coil. Preoperative lymph node staging with imaging modalities (CT, MRI, and DWIBS were compared with the final histological findings. RESULTS: The accuracy of CT, MRI, and DWIBS were 57.7%, 63.5%, and 40.4%. The accuracy of DWIBS with higher sensitivity and negative predictive value for evaluating primary rectal cancer patients was lower than that of CT and MRI. Nodal staging agreement between imaging and pathology was fairly strong for CT and MRI (Kappa value = 0.331 and 0.348, P<0.01 but was relatively weaker for DWIBS (Kappa value = 0.174, P<0.05. The accuracy was 57.7% and 59.6%, respectively, for CT and MRI when the lymph node border information was used as the criteria, and was 57.7% and 61.5%, respectively, for enhanced CT and MRI when the lymph node enhancement pattern was used as the criteria. CONCLUSION: MRI is more accurate than CT in predicting nodal involvement in primary rectal carcinoma patients in the absence of tumor invasion into pelvic structures. DWIBS has a great diagnostic value in differentiating small malignant from benign lymph nodes.

  7. Accuracy and Precision of Three-Dimensional Low Dose CT Compared to Standard RSA in Acetabular Cups: An Experimental Study.

    Science.gov (United States)

    Brodén, Cyrus; Olivecrona, Henrik; Maguire, Gerald Q; Noz, Marilyn E; Zeleznik, Michael P; Sköldenberg, Olof

    2016-01-01

    Background and Purpose. The gold standard for detection of implant wear and migration is currently radiostereometry (RSA). The purpose of this study is to compare a three-dimensional computed tomography technique (3D CT) to standard RSA as an alternative technique for measuring migration of acetabular cups in total hip arthroplasty. Materials and Methods. With tantalum beads, we marked one cemented and one uncemented cup and mounted these on a similarly marked pelvic model. A comparison was made between 3D CT and standard RSA for measuring migration. Twelve repeated stereoradiographs and CT scans with double examinations in each position and gradual migration of the implants were made. Precision and accuracy of the 3D CT were calculated. Results. The accuracy of the 3D CT ranged between 0.07 and 0.32 mm for translations and 0.21 and 0.82° for rotation. The precision ranged between 0.01 and 0.09 mm for translations and 0.06 and 0.29° for rotations, respectively. For standard RSA, the precision ranged between 0.04 and 0.09 mm for translations and 0.08 and 0.32° for rotations, respectively. There was no significant difference in precision between 3D CT and standard RSA. The effective radiation dose of the 3D CT method, comparable to RSA, was estimated to be 0.33 mSv. Interpretation. Low dose 3D CT is a comparable method to standard RSA in an experimental setting.

  8. Improved accuracy of quantitative parameter estimates in dynamic contrast-enhanced CT study with low temporal resolution

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Sun Mo, E-mail: Sunmo.Kim@rmp.uhn.on.ca [Radiation Medicine Program, Princess Margaret Hospital/University Health Network, Toronto, Ontario M5G 2M9 (Canada); Haider, Masoom A. [Department of Medical Imaging, Sunnybrook Health Sciences Centre, Toronto, Ontario M4N 3M5, Canada and Department of Medical Imaging, University of Toronto, Toronto, Ontario M5G 2M9 (Canada); Jaffray, David A. [Radiation Medicine Program, Princess Margaret Hospital/University Health Network, Toronto, Ontario M5G 2M9, Canada and Department of Radiation Oncology, University of Toronto, Toronto, Ontario M5G 2M9 (Canada); Yeung, Ivan W. T. [Radiation Medicine Program, Princess Margaret Hospital/University Health Network, Toronto, Ontario M5G 2M9 (Canada); Department of Medical Physics, Stronach Regional Cancer Centre, Southlake Regional Health Centre, Newmarket, Ontario L3Y 2P9 (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario M5G 2M9 (Canada)

    2016-01-15

    Purpose: A previously proposed method to reduce radiation dose to patient in dynamic contrast-enhanced (DCE) CT is enhanced by principal component analysis (PCA) filtering which improves the signal-to-noise ratio (SNR) of time-concentration curves in the DCE-CT study. The efficacy of the combined method to maintain the accuracy of kinetic parameter estimates at low temporal resolution is investigated with pixel-by-pixel kinetic analysis of DCE-CT data. Methods: The method is based on DCE-CT scanning performed with low temporal resolution to reduce the radiation dose to the patient. The arterial input function (AIF) with high temporal resolution can be generated with a coarsely sampled AIF through a previously published method of AIF estimation. To increase the SNR of time-concentration curves (tissue curves), first, a region-of-interest is segmented into squares composed of 3 × 3 pixels in size. Subsequently, the PCA filtering combined with a fraction of residual information criterion is applied to all the segmented squares for further improvement of their SNRs. The proposed method was applied to each DCE-CT data set of a cohort of 14 patients at varying levels of down-sampling. The kinetic analyses using the modified Tofts’ model and singular value decomposition method, then, were carried out for each of the down-sampling schemes between the intervals from 2 to 15 s. The results were compared with analyses done with the measured data in high temporal resolution (i.e., original scanning frequency) as the reference. Results: The patients’ AIFs were estimated to high accuracy based on the 11 orthonormal bases of arterial impulse responses established in the previous paper. In addition, noise in the images was effectively reduced by using five principal components of the tissue curves for filtering. Kinetic analyses using the proposed method showed superior results compared to those with down-sampling alone; they were able to maintain the accuracy in the

  9. A formulation of tissue- and water-equivalent materials using the stoichiometric analysis method for CT-number calibration in radiotherapy treatment planning

    Science.gov (United States)

    Yohannes, Indra; Kolditz, Daniel; Langner, Oliver; Kalender, Willi A.

    2012-03-01

    Tissue- and water-equivalent materials (TEMs) are widely used in quality assurance and calibration procedures, both in radiodiagnostics and radiotherapy. In radiotherapy, particularly, the TEMs are often used for computed tomography (CT) number calibration in treatment planning systems. However, currently available TEMs may not be very accurate in the determination of the calibration curves due to their limitation in mimicking radiation characteristics of the corresponding real tissues in both low- and high-energy ranges. Therefore, we are proposing a new formulation of TEMs using a stoichiometric analysis method to obtain TEMs for the calibration purposes. We combined the stoichiometric calibration and the basic data method to compose base materials to develop TEMs matching standard real tissues from ICRU Report 44 and 46. First, the CT numbers of six materials with known elemental compositions were measured to get constants for the stoichiometric calibration. The results of the stoichiometric calibration were used together with the basic data method to formulate new TEMs. These new TEMs were scanned to validate their CT numbers. The electron density and the stopping power calibration curves were also generated. The absolute differences of the measured CT numbers of the new TEMs were less than 4 HU for the soft tissues and less than 22 HU for the bone compared to the ICRU real tissues. Furthermore, the calculated relative electron density and electron and proton stopping powers of the new TEMs differed by less than 2% from the corresponding ICRU real tissues. The new TEMs which were formulated using the proposed technique increase the simplicity of the calibration process and preserve the accuracy of the stoichiometric calibration simultaneously.

  10. Evaluation of reduced-dose CT for acute non-traumatic abdominal pain: evaluation of diagnostic accuracy in comparison to standard-dose CT.

    Science.gov (United States)

    Othman, Ahmed E; Bongers, Malte Niklas; Zinsser, Dominik; Schabel, Christoph; Wichmann, Julian L; Arshid, Rami; Notohamiprodjo, Mike; Nikolaou, Konstantin; Bamberg, Fabian

    2018-01-01

    Background Patients with acute non-traumatic abdominal pain often undergo abdominal computed tomography (CT). However, abdominal CT is associated with high radiation exposure. Purpose To evaluate diagnostic performance of a reduced-dose 100 kVp CT protocol with advanced modeled iterative reconstruction as compared to a linearly blended 120 kVp protocol for assessment of acute, non-traumatic abdominal pain. Material and Methods Two radiologists assessed 100 kVp and linearly blended 120 kVp series of 112 consecutive patients with acute non-traumatic pain (onset diagnostic confidence. Both 100 kVp and linearly blended 120 kVp series were quantitatively evaluated regarding radiation dose and image noise. Comparative statistics and diagnostic accuracy was calculated using receiver operating curve (ROC) statistics, with final clinical diagnosis/clinical follow-up as reference standard. Results Image quality was high for both series without detectable significant differences ( P = 0.157). Image noise and artifacts were rated low for both series but significantly higher for 100 kVp ( P ≤ 0.021). Diagnostic accuracy was high for both series (120 kVp: area under the curve [AUC] = 0.950, sensitivity = 0.958, specificity = 0.941; 100 kVp: AUC ≥ 0.910, sensitivity ≥ 0.937, specificity = 0.882; P ≥ 0.516) with almost perfect inter-rater agreement (Kappa = 0.939). Diagnostic confidence was high for both dose levels without significant differences (100 kVp 5, range 4-5; 120 kVp 5, range 3-5; P = 0.134). The 100 kVp series yielded 26.1% lower radiation dose compared with the 120 kVp series (5.72 ± 2.23 mSv versus 7.75 ± 3.02 mSv, P diagnostic accuracy for the assessment of acute non-traumatic abdominal pain.

  11. Radiographic markers for measuring tibial rotation based on CT-reconstructed radiographs. An accuracy and feasibility study

    International Nuclear Information System (INIS)

    Hakimian, David; Khoury, Amal; Mosheiff, Rami; Liebergall, Meir; Weil, Yoram A.

    2018-01-01

    Malreduction in the axial plane (malrotation) following tibial fracture surgery is often undiagnosed. A few clinical and radiographic methods have been proposed for measuring tibial rotation intraoperatively, yet have failed to match the accuracy of computed tomography (CT). The aim of this study was to develop radiographic tools for future intraoperative assessment of the tibial shaft rotation profile. The setting was a laboratory computerized analysis. Twenty lower limb CT scans were used to construct a three-dimensional (3D) model using AMIRA copyright software. A virtual 3D cylinder was implanted in the posterior condylar line and in the transmalleolar axis. The 3D models were used to simulate four standard knee and ankle plain radiographs. On each radiograph, four landmarks were depicted by two observers and their relation with the cylinder was measured and analyzed for accuracy and reproducibility. A cadaveric lower leg was implanted with two Kirschner wires. A CT scan was performed in addition to 2D fluoroscopy. The simulated radiographs and the fluoroscopy were compared for accuracy. Measurement of the landmarks showed reliability in most of the knee anteroposterior and ankle mortise radiographs (coefficients of variation < 0.01 and = 0.01) respectively. Cadaveric measurement of the landmarks using real fluoroscopy and simulated radiographs were similar. To date, no reliable and common methods have been reported for the evaluation of tibial axial rotation. We propose a model in which simple radiographic landmarks can be used to calculate a 3D coordinate system that accurately assesses the axial rotation angle of the tibial shaft. (orig.)

  12. Radiographic markers for measuring tibial rotation based on CT-reconstructed radiographs. An accuracy and feasibility study

    Energy Technology Data Exchange (ETDEWEB)

    Hakimian, David; Khoury, Amal; Mosheiff, Rami; Liebergall, Meir; Weil, Yoram A. [Hadassah Hebrew University Hospital, Department of Orthopaedics, Jerusalem (Israel)

    2018-04-15

    Malreduction in the axial plane (malrotation) following tibial fracture surgery is often undiagnosed. A few clinical and radiographic methods have been proposed for measuring tibial rotation intraoperatively, yet have failed to match the accuracy of computed tomography (CT). The aim of this study was to develop radiographic tools for future intraoperative assessment of the tibial shaft rotation profile. The setting was a laboratory computerized analysis. Twenty lower limb CT scans were used to construct a three-dimensional (3D) model using AMIRA copyright software. A virtual 3D cylinder was implanted in the posterior condylar line and in the transmalleolar axis. The 3D models were used to simulate four standard knee and ankle plain radiographs. On each radiograph, four landmarks were depicted by two observers and their relation with the cylinder was measured and analyzed for accuracy and reproducibility. A cadaveric lower leg was implanted with two Kirschner wires. A CT scan was performed in addition to 2D fluoroscopy. The simulated radiographs and the fluoroscopy were compared for accuracy. Measurement of the landmarks showed reliability in most of the knee anteroposterior and ankle mortise radiographs (coefficients of variation < 0.01 and = 0.01) respectively. Cadaveric measurement of the landmarks using real fluoroscopy and simulated radiographs were similar. To date, no reliable and common methods have been reported for the evaluation of tibial axial rotation. We propose a model in which simple radiographic landmarks can be used to calculate a 3D coordinate system that accurately assesses the axial rotation angle of the tibial shaft. (orig.)

  13. [18F]Fluorodeoxyglucose (FDG)-Positron Emission Tomography (PET)/Computed Tomography (CT) in Suspected Recurrent Breast Cancer: A Prospective Comparative Study of Dual-Time-Point FDG-PET/CT, Contrast-Enhanced CT, and Bone Scintigraphy.

    Science.gov (United States)

    Hildebrandt, Malene Grubbe; Gerke, Oke; Baun, Christina; Falch, Kirsten; Hansen, Jeanette Ansholm; Farahani, Ziba Ahangarani; Petersen, Henrik; Larsen, Lisbet Brønsro; Duvnjak, Sandra; Buskevica, Inguna; Bektas, Selma; Søe, Katrine; Jylling, Anne Marie Bak; Ewertz, Marianne; Alavi, Abass; Høilund-Carlsen, Poul Flemming

    2016-06-01

    To prospectively investigate the diagnostic accuracy of [(18)F]fluorodeoxyglucose (FDG)-positron emission tomography (PET)/computed tomography (CT) with dual-time-point imaging, contrast-enhanced CT (ceCT), and bone scintigraphy (BS) in patients with suspected breast cancer recurrence. One hundred women with suspected recurrence of breast cancer underwent 1-hour and 3-hour FDG-PET/CT, ceCT, and BS within approximately 10 days. The study was powered to estimate the precision of the individual imaging tests. Images were visually interpreted using a four-point assessment scale, and readers were blinded to other test results. The reference standard was biopsy along with treatment decisions and clinical follow-up (median, 17 months). FDG-PET/CT resulted in no false negatives and fewer false positives than the other imaging techniques. Accuracy of results were similar for 1-hour and 3-hour FDG-PET/CT. For distant recurrence, the area under the receiver operating curve was 0.99 (95% CI, 0.97 to 1) for FDG-PET/CT, 0.84 (95% CI, 0.73 to 0.94) for ceCT, and 0.86 (95% CI, 0.77 to 0.94) for the combined ceCT+BS. Of 100 patients, 22 (22%) were verified with distant recurrence, and 18 of these had bone involvement. Nineteen patients (19%) had local recurrence only. In exploratory analyses, diagnostic accuracy of FDG-PET/CT was better than ceCT alone or ceCT combined with BS in diagnosing distant, bone, and local recurrence, shown by a greater area under the receiver operating curve and higher sensitivity, specificity, and superior likelihood ratios. FDG-PET/CT was accurate in diagnosing recurrence in breast cancer patients. It allowed for distant recurrence to be correctly ruled out and resulted in only a small number of false-positive cases. Exploratory findings suggest that FDG-PET/CT has greater accuracy than conventional imaging technologies in this patient group. © 2016 by American Society of Clinical Oncology.

  14. Diagnostic accuracy of 64-slice multidetector CT for detection of in-stent restenosis in an unselected, consecutive patient population

    Energy Technology Data Exchange (ETDEWEB)

    Haraldsdottir, Sigurdis, E-mail: sigurdisha@gmail.com [Boston Medical Center, 72 East Concord Street (Evans 124), Boston, MA, 02118 (United States); Gudnason, Thorarinn, E-mail: thorgudn@landspitali.is [Landspitali University Hospital, Hringbraut, 101 Reykjavik (Iceland); Sigurdsson, Axel F., E-mail: axelfsig@landspitali.is [Landspitali University Hospital, Hringbraut, 101 Reykjavik (Iceland); Gudjonsdottir, Jonina, E-mail: jonina@rd.is [Rontgen Domus Medica, Egilsgata 3, 101 Reykjavik (Iceland); Lehman, Sam J., E-mail: slehman@partners.org [Massachusetts General Hospital, 165 Cambridge Street, Suite 400, Boston, MA 02114 (United States); Eyjolfsson, Kristjan, E-mail: kristey@landspitali.is [Landspitali University Hospital, Hringbraut, 101 Reykjavik (Iceland); Scheving, Sigurpall S., E-mail: sigurpal@landspitali.is [Landspitali University Hospital, Hringbraut, 101 Reykjavik (Iceland); Gibson, C. Michael, E-mail: mgibson@perfuse.org [Beth Israel Deaconess Medical Center, 330 Brookline Avenue, Boston, MA 02115 (United States); Hoffmann, Udo, E-mail: uhoffmann@partners.org [Massachusetts General Hospital, 165 Cambridge Street, Suite 400, Boston, MA 02114 (United States); Jonsdottir, Birna, E-mail: birna@rd.is [Rontgen Domus Medica, Egilsgata 3, 101 Reykjavik (Iceland); Andersen, Karl, E-mail: andersen@landspitali.is [Landspitali University Hospital, Hringbraut, 101 Reykjavik (Iceland)

    2010-11-15

    Objectives: To investigate the diagnostic accuracy of 64-slice multidetector computed tomography (64-CT) for detection of in-stent restenosis (ISR) in an unselected, consecutive patient population. Background: Detection of in-stent restenosis by cardiac CT would be a major advance for the evaluation of patients suspected of having ISR. However, the diagnostic accuracy of current generation 64-CT in this context is not fully established. Methods: We conducted a prospective study on patients with stable angina or acute coronary syndrome with no prior history of coronary artery disease. Six months after percutaneous coronary intervention (PCI) with stent placement they underwent a 64-CT scan (Toshiba Multi-Slice Aquilion 64) and consequently a repeat coronary angiography for comparison. Cardiac CT data sets were analyzed for the presence of in-stent restenosis by two independent expert readers blinded to the coronary angiographic data. Results: Ninety-three patients with a total of 140 stents were evaluated. Males comprised 82% of the study group and the mean age was 63 {+-} 10 years. The mean time from PCI to the repeat coronary angiography was 208 {+-} 37 days and the mean time from 64-CT to repeat coronary angiography was 3.7 {+-} 4.9 days. The restenosis rate according to coronary angiography was 26%. Stent diameter, strut thickness, heart rate and body mass index (BMI) significantly affected image quality. The sensitivity, specificity, positive and negative predictive values of 64-CT for detection of in-stent restenosis were 27%, 95%, 67% and 78%, respectively. Conclusions: Current generation, 64-slice CT, remains limited in its ability to accurately detect in-stent restenosis.

  15. SU-E-T-396: Dosimetric Accuracy of Proton Therapy for Patients with Metal Implants in CT Scans Using Metal Deletion Technique (MDT) Artifacts Reduction

    International Nuclear Information System (INIS)

    Li, X; Kantor, M; Zhu, X; Frank, S; Sahoo, N; Li, H

    2014-01-01

    Purpose: To evaluate the dosimetric accuracy for proton therapy patients with metal implants in CT using metal deletion technique (MDT) artifacts reduction. Methods: Proton dose accuracies under CT metal artifacts were first evaluated using a water phantom with cylindrical inserts of different materials (titanium and steel). Ranges and dose profiles along different beam angles were calculated using treatment planning system (Eclipse version 8.9) on uncorrected CT, MDT CT, and manually-corrected CT, where true Hounsfield units (water) were assigned to the streak artifacts. In patient studies, the treatment plans were developed on manually-corrected CTs, then recalculated on MDT and uncorrected CTs. DVH indices were compared between the dose distributions on all the CTs. Results: For water phantom study with 1/2 inch titanium insert, the proton range differences estimated by MDT CT were with 1% for all beam angles, while the range error can be up to 2.6% for uncorrected CT. For the study with 1 inch stainless steel insert, the maximum range error calculated by MDT CT was 1.09% among all the beam angles compared with maximum range error with 4.7% for uncorrected CT. The dose profiles calculated on MDT CTs for both titanium and steel inserts showed very good agreements with the ones calculated on manually-corrected CTs, while large dose discrepancies calculated using uncorrected CTs were observed in the distal end region of the proton beam. The patient study showed similar dose distribution and DVHs for organs near the metal artifacts recalculated on MDT CT compared with the ones calculated on manually-corrected CT, while the differences between uncorrected and corrected CTs were much pronounced. Conclusion: In proton therapy, large dose error could occur due to metal artifact. The MDT CT can be used for proton dose calculation to achieve similar dose accuracy as the current clinical practice using manual correction

  16. Is diagnostic accuracy for detecting pulmonary nodules in chest CT reduced after a long day of reading?

    Science.gov (United States)

    Krupinski, Elizabeth A.; Berbaum, Kevin S.; Caldwell, Robert; Schartz, Kevin M.

    2012-02-01

    Radiologists are reading more cases with more images, especially in CT and MRI and thus working longer hours than ever before. There have been concerns raised regarding fatigue and whether it impacts diagnostic accuracy. This study measured the impact of reader visual fatigue by assessing symptoms, visual strain via dark focus of accommodation, and diagnostic accuracy. Twenty radiologists and 20 radiology residents were given two diagnostic performance tests searching CT chest sequences for a solitary pulmonary nodule before (rested) and after (tired) a day of clinical reading. 10 cases used free search and navigation, and the other 100 cases used preset scrolling speed and duration. Subjects filled out the Swedish Occupational Fatigue Inventory (SOFI) and the oculomotor strain subscale of the Simulator Sickness Questionnaire (SSQ) before each session. Accuracy was measured using ROC techniques. Using Swensson's technique yields an ROC area = 0.86 rested vs. 0.83 tired, p (one-tailed) = 0.09. Using Swensson's LROC technique yields an area = 0.73 rested vs. 0.66 tired, p (one-tailed) = 0.09. Using Swensson's Loc Accuracy technique yields an area = 0.77 rested vs. 0.72 tired, p (one-tailed) = 0.13). Subjective measures of fatigue increased significantly from early to late reading. To date, the results support our findings with static images and detection of bone fractures. Radiologists at the end of a long work day experience greater levels of measurable visual fatigue or strain, contributing to a decrease in diagnostic accuracy. The decrease in accuracy was not as great however as with static images.

  17. Diagnostic accuracy at several reduced radiation dose levels for CT imaging in the diagnosis of appendicitis

    Science.gov (United States)

    Zhang, Di; Khatonabadi, Maryam; Kim, Hyun; Jude, Matilda; Zaragoza, Edward; Lee, Margaret; Patel, Maitraya; Poon, Cheryce; Douek, Michael; Andrews-Tang, Denise; Doepke, Laura; McNitt-Gray, Shawn; Cagnon, Chris; DeMarco, John; McNitt-Gray, Michael

    2012-03-01

    Purpose: While several studies have investigated the tradeoffs between radiation dose and image quality (noise) in CT imaging, the purpose of this study was to take this analysis a step further by investigating the tradeoffs between patient radiation dose (including organ dose) and diagnostic accuracy in diagnosis of appendicitis using CT. Methods: This study was IRB approved and utilized data from 20 patients who underwent clinical CT exams for indications of appendicitis. Medical record review established true diagnosis of appendicitis, with 10 positives and 10 negatives. A validated software tool used raw projection data from each scan to create simulated images at lower dose levels (70%, 50%, 30%, 20% of original). An observer study was performed with 6 radiologists reviewing each case at each dose level in random order over several sessions. Readers assessed image quality and provided confidence in their diagnosis of appendicitis, each on a 5 point scale. Liver doses at each case and each dose level were estimated using Monte Carlo simulation based methods. Results: Overall diagnostic accuracy varies across dose levels: 92%, 93%, 91%, 90% and 90% across the 100%, 70%, 50%, 30% and 20% dose levels respectively. And it is 93%, 95%, 88%, 90% and 90% across the 13.5-22mGy, 9.6-13.5mGy, 6.4-9.6mGy, 4-6.4mGy, and 2-4mGy liver dose ranges respectively. Only 4 out of 600 observations were rated "unacceptable" for image quality. Conclusion: The results from this pilot study indicate that the diagnostic accuracy does not change dramatically even at significantly reduced radiation dose.

  18. Calibration of megavoltage cone-beam CT for radiotherapy dose calculations: Correction of cupping artifacts and conversion of CT numbers to electron density

    International Nuclear Information System (INIS)

    Petit, Steven F.; Elmpt, Wouter J. C. van; Nijsten, Sebastiaan M. J. J. G.; Lambin, Philippe; Dekker, Andre L. A. J.

    2008-01-01

    Megavoltage cone-beam CT (MV CBCT) is used for three-dimensional imaging of the patient anatomy on the treatment table prior to or just after radiotherapy treatment. To use MV CBCT images for radiotherapy dose calculation purposes, reliable electron density (ED) distributions are needed. Patient scatter, beam hardening and softening effects result in cupping artifacts in MV CBCT images and distort the CT number to ED conversion. A method based on transmission images is presented to correct for these effects without using prior knowledge of the object's geometry. The scatter distribution originating from the patient is calculated with pencil beam scatter kernels that are fitted based on transmission measurements. The radiological thickness is extracted from the scatter subtracted transmission images and is then converted to the primary transmission used in the cone-beam reconstruction. These corrections are performed in an iterative manner, without using prior knowledge regarding the geometry and composition of the object. The method was tested using various homogeneous and inhomogeneous phantoms with varying shapes and compositions, including a phantom with different electron density inserts, phantoms with large density variations, and an anthropomorphic head phantom. For all phantoms, the cupping artifact was substantially removed from the images and a linear relation between the CT number and electron density was found. After correction the deviations in reconstructed ED from the true values were reduced from up to 0.30 ED units to 0.03 for the majority of the phantoms; the residual difference is equal to the amount of noise in the images. The ED distributions were evaluated in terms of absolute dose calculation accuracy for homogeneous cylinders of different size; errors decreased from 7% to below 1% in the center of the objects for the uncorrected and corrected images, respectively, and maximum differences were reduced from 17% to 2%, respectively. The

  19. Accuracy of volume measurement using 3D ultrasound and development of CT-3D US image fusion algorithm for prostate cancer radiotherapy

    International Nuclear Information System (INIS)

    Baek, Jihye; Huh, Jangyoung; Hyun An, So; Oh, Yoonjin; Kim, Myungsoo; Kim, DongYoung; Chung, Kwangzoo; Cho, Sungho; Lee, Rena

    2013-01-01

    Purpose: To evaluate the accuracy of measuring volumes using three-dimensional ultrasound (3D US), and to verify the feasibility of the replacement of CT-MR fusion images with CT-3D US in radiotherapy treatment planning. Methods: Phantoms, consisting of water, contrast agent, and agarose, were manufactured. The volume was measured using 3D US, CT, and MR devices. A CT-3D US and MR-3D US image fusion software was developed using the Insight Toolkit library in order to acquire three-dimensional fusion images. The quality of the image fusion was evaluated using metric value and fusion images. Results: Volume measurement, using 3D US, shows a 2.8 ± 1.5% error, 4.4 ± 3.0% error for CT, and 3.1 ± 2.0% error for MR. The results imply that volume measurement using the 3D US devices has a similar accuracy level to that of CT and MR. Three-dimensional image fusion of CT-3D US and MR-3D US was successfully performed using phantom images. Moreover, MR-3D US image fusion was performed using human bladder images. Conclusions: 3D US could be used in the volume measurement of human bladders and prostates. CT-3D US image fusion could be used in monitoring the target position in each fraction of external beam radiation therapy. Moreover, the feasibility of replacing the CT-MR image fusion to the CT-3D US in radiotherapy treatment planning was verified.

  20. Sensitivity and accuracy of volumetry of pulmonary nodules on low-dose 16- and 64-row multi-detector CT : an anthropomorphic phantom study

    NARCIS (Netherlands)

    Xie, X.; Zhao, Yingru; Snijder, R.A.; van Ooijen, P.M.; de Jong, P.A.; Oudkerk, M.; de Bock, G.H.; Vliegenthart, R.; Greuter, M.J.

    To assess the sensitivity of detection and accuracy of volumetry by manual and semi-automated quantification of artificial pulmonary nodules in an anthropomorphic thoracic phantom on low-dose CT. Fifteen artificial spherical nodules (diameter 3, 5, 8, 10 and 12 mm; CT densities -800, -630 and +100

  1. Comparison of the diagnostic accuracy of PET/MRI to PET/CT-acquired FDG brain exams for seizure focus detection: a prospective study

    Energy Technology Data Exchange (ETDEWEB)

    Paldino, Michael J.; Jones, Jeremy Y.; Mahmood, Nadia; Sher, Andrew; Hayatghaibi, Shireen; Seghers, Victor [Texas Children' s Hospital, Department of Radiology, Houston, TX (United States); Yang, Erica [SimonMed Imaging, Department of Radiology, Scottsdale, AZ (United States); Zhang, Wei [Texas Children' s Hospital, Outcomes and Impact Service, Houston, TX (United States); Krishnamurthy, Ramkumar [Nationwide Children' s Hospital, Department of Radiology, Columbus, OH (United States)

    2017-10-15

    There is great interest in positron emission tomography (PET)/magnetic resonance (MR) as a clinical tool due to its capacity to provide diverse diagnostic information in a single exam. The goal of this exam is to compare the diagnostic accuracy of PET/MR-acquired [F-18]2-fluoro-2-deoxyglucose (FDG) brain exams to that of PET/CT with respect to identifying seizure foci in children with localization-related epilepsy. Institutional Review Board approval and informed consent were obtained for this Health Insurance Portability and Accountability Act-compliant, prospective study. All patients referred for clinical FDG-PET/CT exams of the brain at our institution for a diagnosis of localization-related epilepsy were prospectively recruited to undergo an additional FDG-PET acquisition on a tandem PET/MR system. Attenuation-corrected FDG images acquired at PET/MR and PET/CT were interpreted independently by five expert readers. Readers were blinded to the scanner used for acquisition and attenuation correction as well as all other clinical and imaging data. A Likert scale scoring system (1-5) was used to assess image quality. The locale of seizure origin determined at multidisciplinary epilepsy surgery work rounds was considered the reference standard. Non-inferiority testing for paired data was used to compare the diagnostic accuracy of PET/MR to that of PET/CT. The final study population comprised 35 patients referred for a diagnosis of localization-related epilepsy (age range: 2-19 years; median: 11 years; 21 males, 14 females). Image quality did not differ significantly between the two modalities. The accuracy of PET/MR was not inferior to that of PET/CT for localization of a seizure focus (P=0.017). The diagnostic accuracy of FDG-PET images acquired on a PET/MR scanner and generated using MR-based attenuation correction was not inferior to that of PET images processed by traditional CT-based correction. (orig.)

  2. Comparison of the diagnostic accuracy of PET/MRI to PET/CT-acquired FDG brain exams for seizure focus detection: a prospective study

    International Nuclear Information System (INIS)

    Paldino, Michael J.; Jones, Jeremy Y.; Mahmood, Nadia; Sher, Andrew; Hayatghaibi, Shireen; Seghers, Victor; Yang, Erica; Zhang, Wei; Krishnamurthy, Ramkumar

    2017-01-01

    There is great interest in positron emission tomography (PET)/magnetic resonance (MR) as a clinical tool due to its capacity to provide diverse diagnostic information in a single exam. The goal of this exam is to compare the diagnostic accuracy of PET/MR-acquired [F-18]2-fluoro-2-deoxyglucose (FDG) brain exams to that of PET/CT with respect to identifying seizure foci in children with localization-related epilepsy. Institutional Review Board approval and informed consent were obtained for this Health Insurance Portability and Accountability Act-compliant, prospective study. All patients referred for clinical FDG-PET/CT exams of the brain at our institution for a diagnosis of localization-related epilepsy were prospectively recruited to undergo an additional FDG-PET acquisition on a tandem PET/MR system. Attenuation-corrected FDG images acquired at PET/MR and PET/CT were interpreted independently by five expert readers. Readers were blinded to the scanner used for acquisition and attenuation correction as well as all other clinical and imaging data. A Likert scale scoring system (1-5) was used to assess image quality. The locale of seizure origin determined at multidisciplinary epilepsy surgery work rounds was considered the reference standard. Non-inferiority testing for paired data was used to compare the diagnostic accuracy of PET/MR to that of PET/CT. The final study population comprised 35 patients referred for a diagnosis of localization-related epilepsy (age range: 2-19 years; median: 11 years; 21 males, 14 females). Image quality did not differ significantly between the two modalities. The accuracy of PET/MR was not inferior to that of PET/CT for localization of a seizure focus (P=0.017). The diagnostic accuracy of FDG-PET images acquired on a PET/MR scanner and generated using MR-based attenuation correction was not inferior to that of PET images processed by traditional CT-based correction. (orig.)

  3. Alignment of CT images of skull dysmorphology using anatomy-based perpendicular axes

    International Nuclear Information System (INIS)

    Yoo, Sun K; Kim, Yong O; Kim, Hee-Joung; Kim, Nam H; Jang, Young Beom; Kim, Kee-Deog; Lee, Hye-Yeon

    2003-01-01

    Rigid body registration of 3D CT scans, based on manual identification of homologous landmarks, is useful for the visual analysis of skull dysmorphology. In this paper, a robust and simple alignment method was proposed to allow for the comparison of skull morphologies, within and between individuals with craniofacial anomalies, based on 3D CT scans, and the minimum number of anatomical landmarks, under rigidity and uniqueness constraints. Three perpendicular axes, extracted from anatomical landmarks, define the absolute coordinate system, through a rigid body transformation, to align multiple CT images for different patients and acquisition times. The accuracy of the alignment method depends on the accuracy of the localized landmarks and target points. The numerical simulation generalizes the accuracy requirements of the alignment method. Experiments using a human dried skull specimen, and ten sets of skull CT images (the pre- and post-operative CT scans of four plagiocephaly, and one fibrous dysplasia patients), demonstrated the feasibility of the technique in clinical practice

  4. Accuracy of radiotherapy dose calculations based on cone-beam CT: comparison of deformable registration and image correction based methods

    Science.gov (United States)

    Marchant, T. E.; Joshi, K. D.; Moore, C. J.

    2018-03-01

    Radiotherapy dose calculations based on cone-beam CT (CBCT) images can be inaccurate due to unreliable Hounsfield units (HU) in the CBCT. Deformable image registration of planning CT images to CBCT, and direct correction of CBCT image values are two methods proposed to allow heterogeneity corrected dose calculations based on CBCT. In this paper we compare the accuracy and robustness of these two approaches. CBCT images for 44 patients were used including pelvis, lung and head & neck sites. CBCT HU were corrected using a ‘shading correction’ algorithm and via deformable registration of planning CT to CBCT using either Elastix or Niftyreg. Radiotherapy dose distributions were re-calculated with heterogeneity correction based on the corrected CBCT and several relevant dose metrics for target and OAR volumes were calculated. Accuracy of CBCT based dose metrics was determined using an ‘override ratio’ method where the ratio of the dose metric to that calculated on a bulk-density assigned version of the same image is assumed to be constant for each patient, allowing comparison to the patient’s planning CT as a gold standard. Similar performance is achieved by shading corrected CBCT and both deformable registration algorithms, with mean and standard deviation of dose metric error less than 1% for all sites studied. For lung images, use of deformed CT leads to slightly larger standard deviation of dose metric error than shading corrected CBCT with more dose metric errors greater than 2% observed (7% versus 1%).

  5. Systematic radiation dose optimization of abdominal dual-energy CT on a second-generation dual-source CT scanner: assessment of the accuracy of iodine uptake measurement and image quality in an in vitro and in vivo investigations.

    Science.gov (United States)

    Schindera, Sebastian T; Zaehringer, Caroline; D'Errico, Luigia; Schwartz, Fides; Kekelidze, Maka; Szucs-Farkas, Zsolt; Benz, Matthias R

    2017-10-01

    To assess the accuracy of iodine quantification in a phantom study at different radiation dose levels with dual-energy dual-source CT and to evaluate image quality and radiation doses in patients undergoing a single-energy and two dual-energy abdominal CT protocols. In a phantom study, the accuracy of iodine quantification (4.5-23.5 mgI/mL) was evaluated using the manufacturer-recommended and three dose-optimized dual-energy protocols. In a patient study, 75 abdomino-pelvic CT examinations were acquired as follows: 25 CT scans with the manufacturer-recommended dual-energy protocol (protocol A); 25 CT scans with a dose-optimized dual-energy protocol (protocol B); and 25 CT scans with a single-energy CT protocol (protocol C). CTDI vol and objective noise were measured. Five readers scored each scan according to six subjective image quality parameters (noise, contrast, artifacts, visibility of small structures, sharpness, overall diagnostic confidence). In the phantom study, differences between the real and measured iodine concentrations ranged from -8.8% to 17.0% for the manufacturer-recommended protocol and from -1.6% to 20.5% for three dose-optimized protocols. In the patient study, the CTDI vol of protocol A, B, and C were 12.5 ± 1.9, 7.5 ± 1.2, and 6.5 ± 1.7 mGycm, respectively (p dual-energy and the single-energy protocol. A dose reduction of 41% is feasible for the manufacturer-recommended, abdominal dual-energy CT protocol, as it maintained the accuracy of iodine measurements and subjective image quality compared to a single-energy protocol.

  6. Spatial Component Position in Total Hip Arthroplasty. Accuracy and repeatability with a new CT method

    International Nuclear Information System (INIS)

    Olivecrona, H.

    2003-01-01

    Purpose: 3D detection of centerpoints of prosthetic cup and head after total hip arthroplasty (THA) using CT. Material and Methods: Two CT examinations, 10 min apart, were obtained from each of 10 patients after THA. Two independent examiners placed landmarks in images of the prosthetic cup and head. All landmarking was repeated after 1 week. Centerpoints were calculated and compared. Results: Within volumes, all measurements of centerpoints of cup and head fell, with a 95% confidence, within one CT-voxel of any other measurement of the same object. Across two volumes, the mean error of distance between center of cup and prosthetic head was 1.4 mm (SD 0.73). Intra- and interobserver 95% accuracy limit was below 2 mm within and below 3 mm across volumes. No difference between intra- and interobserver measurements occurred. A formula for converting finite sets of point landmarks in the radiolucent tread of the cup to a centerpoint was stable. The percent difference of the landmark distances from a calculated spherical surface was within one CT-voxel. This data was normally distributed and not dependent on observer or trial. Conclusion: The true 3D position of the centers of cup and prosthetic head can be detected using CT. Spatial relationship between the components can be analyzed visually and numerically

  7. Spatial Component Position in Total Hip Arthroplasty. Accuracy and repeatability with a new CT method

    Energy Technology Data Exchange (ETDEWEB)

    Olivecrona, H. [Soedersjukhuset, Stockholm (Sweden). Dept. of Hand Surgery; Weidenhielm, L. [Karolinska Hospital, Stockholm (Sweden). Dept. of Orthopedics; Olivecrona, L. [Karolinska Hospital, Stockholm (Sweden). Dept. of Radiology; Noz, M.E. [New York Univ. School of Medicine, NY (United States). Dept. of Radiology; Maguire, G.Q. [Royal Inst. of Tech., Kista (Sweden). Inst. for Microelectronics and Information Technology; Zeleznik, M. P. [Univ. of Utah, Salt Lake City, UT (United States). Dept. of Radiation Oncology; Svensson, L. [Royal Inst. of Tech., Stockholm (Sweden). Dept. of Mathematics; Jonson, T. [Eskadern Foeretagsutveckling AB, Goeteborg (Sweden)

    2003-03-01

    Purpose: 3D detection of centerpoints of prosthetic cup and head after total hip arthroplasty (THA) using CT. Material and Methods: Two CT examinations, 10 min apart, were obtained from each of 10 patients after THA. Two independent examiners placed landmarks in images of the prosthetic cup and head. All landmarking was repeated after 1 week. Centerpoints were calculated and compared. Results: Within volumes, all measurements of centerpoints of cup and head fell, with a 95% confidence, within one CT-voxel of any other measurement of the same object. Across two volumes, the mean error of distance between center of cup and prosthetic head was 1.4 mm (SD 0.73). Intra- and interobserver 95% accuracy limit was below 2 mm within and below 3 mm across volumes. No difference between intra- and interobserver measurements occurred. A formula for converting finite sets of point landmarks in the radiolucent tread of the cup to a centerpoint was stable. The percent difference of the landmark distances from a calculated spherical surface was within one CT-voxel. This data was normally distributed and not dependent on observer or trial. Conclusion: The true 3D position of the centers of cup and prosthetic head can be detected using CT. Spatial relationship between the components can be analyzed visually and numerically.

  8. Dosimetric comparison of stopping power calibration with dual-energy CT and single-energy CT in proton therapy treatment planning

    Energy Technology Data Exchange (ETDEWEB)

    Zhu, Jiahua [Department of Physics, University of Adelaide, Adelaide, SA 5005 (Australia); Penfold, Scott N., E-mail: scott.penfold@adelaide.edu.au [Department of Physics, University of Adelaide, Adelaide, SA 5005, Australia and Department of Medical Physics, Royal Adelaide Hospital, Adelaide, SA 5000 (Australia)

    2016-06-15

    Purpose: The accuracy of proton dose calculation is dependent on the ability to correctly characterize patient tissues with medical imaging. The most common method is to correlate computed tomography (CT) numbers obtained via single-energy CT (SECT) with proton stopping power ratio (SPR). CT numbers, however, cannot discriminate between a change in mass density and change in chemical composition of patient tissues. This limitation can have consequences on SPR calibration accuracy. Dual-energy CT (DECT) is receiving increasing interest as an alternative imaging modality for proton therapy treatment planning due to its ability to discriminate between changes in patient density and chemical composition. In the current work we use a phantom of known composition to demonstrate the dosimetric advantages of proton therapy treatment planning with DECT over SECT. Methods: A phantom of known composition was scanned with a clinical SECT radiotherapy CT-simulator. The phantom was rescanned at a lower X-ray tube potential to generate a complimentary DECT image set. A set of reference materials similar in composition to the phantom was used to perform a stoichiometric calibration of SECT CT number to proton SPRs. The same set of reference materials was used to perform a DECT stoichiometric calibration based on effective atomic number. The known composition of the phantom was used to assess the accuracy of SPR calibration with SECT and DECT. Intensity modulated proton therapy (IMPT) treatment plans were generated with the SECT and DECT image sets to assess the dosimetric effect of the imaging modality. Isodose difference maps and root mean square (RMS) error calculations were used to assess dose calculation accuracy. Results: SPR calculation accuracy was found to be superior, on average, with DECT relative to SECT. Maximum errors of 12.8% and 2.2% were found for SECT and DECT, respectively. Qualitative examination of dose difference maps clearly showed the dosimetric advantages

  9. Accuracy of portable chest X-ray film in detection of intrathoracic abnormal findings in patients after esophagectomy in comparison with CT

    International Nuclear Information System (INIS)

    Shimada, Muneaki; Takeda, Junzo; Ochiai, Ryoichi; Fukushima, Kazuaki; Kouda, Eiichi

    1992-01-01

    To determine an accuracy of portable chest radiographs (CXR) in the diagnosis of pneumothorax, consolidation, passive collapse, subsegmental atelectasis, left and right pleural effusion, findings of 35 pairs of chest CT (CT) and CXR were compared in 21 patients after esophagectomy. The presence of abnormal findings on CXR and CT was evaluated separately by radiologists. Assuming that the diagnoses through CT were correct, sensitivity, specificity, positive predictive value and negative predictive value for CXR were examined in each of 6 abnormal findings. Comparison of CT findings with those of CXR resulted in the following sensitivities (SN) and specificities (SP): pneumothorax: SN=0.17, SP=1.0; consolidation: SN=0.82, SP=1.0; passive collapse: SN=0.83, SP=0.58; subsegmental atelectasis: SN=0.61, SP=1.0; left pleural effusion: SN=0.79, SP=0.29; right pleural effusion: SN=0.9, SP=0.8. These results demonstrated that CXR might miss pneumothorax and overestimate left pleural effusion. Positive predictive value was over 0.79 in each of 6 abnormal findings, which indicated an accuracy of abnormal findings seen on CXR was high. However, negative predictive value in left pleural effusion, which was 0.25, indicated difficulty in correct estimation of left pleural effusion. (author)

  10. Accuracy of iodine quantification using dual energy CT in latest generation dual source and dual layer CT

    Energy Technology Data Exchange (ETDEWEB)

    Pelgrim, Gert Jan; Oudkerk, Matthijs [University of Groningen, University Medical Center Groningen, Center for Medical Imaging - North East Netherlands, P.O. Box EB44, Groningen (Netherlands); Hamersvelt, Robbert W. van; Willemink, Martin J.; Schilham, Arnold; Leiner, Tim [Utrecht University Medical Center, Department of Radiology, Utrecht (Netherlands); Schmidt, Bernhard T.; Flohr, Thomas [Siemens Healthcare GmbH, Forchheim (Germany); Milles, Julien [Philips Healthcare, Best (Netherlands); Vliegenthart, Rozemarijn [University of Groningen, University Medical Center Groningen, Center for Medical Imaging - North East Netherlands, P.O. Box EB44, Groningen (Netherlands); University of Groningen, University Medical Center Groningen, Department of Radiology, Groningen (Netherlands)

    2017-09-15

    To determine the accuracy of iodine quantification with dual energy computed tomography (DECT) in two high-end CT systems with different spectral imaging techniques. Five tubes with different iodine concentrations (0, 5, 10, 15, 20 mg/ml) were analysed in an anthropomorphic thoracic phantom. Adding two phantom rings simulated increased patient size. For third-generation dual source CT (DSCT), tube voltage combinations of 150Sn and 70, 80, 90, 100 kVp were analysed. For dual layer CT (DLCT), 120 and 140 kVp were used. Scans were repeated three times. Median normalized values and interquartile ranges (IQRs) were calculated for all kVp settings and phantom sizes. Correlation between measured and known iodine concentrations was excellent for both systems (R = 0.999-1.000, p < 0.0001). For DSCT, median measurement errors ranged from -0.5% (IQR -2.0, 2.0%) at 150Sn/70 kVp and -2.3% (IQR -4.0, -0.1%) at 150Sn/80 kVp to -4.0% (IQR -6.0, -2.8%) at 150Sn/90 kVp. For DLCT, median measurement errors ranged from -3.3% (IQR -4.9, -1.5%) at 140 kVp to -4.6% (IQR -6.0, -3.6%) at 120 kVp. Larger phantom sizes increased variability of iodine measurements (p < 0.05). Iodine concentration can be accurately quantified with state-of-the-art DECT systems from two vendors. The lowest absolute errors were found for DSCT using the 150Sn/70 kVp or 150Sn/80 kVp combinations, which was slightly more accurate than 140 kVp in DLCT. (orig.)

  11. Non-enhanced CT versus contrast-enhanced CT in integrated PET/CT studies for nodal staging of rectal cancer

    International Nuclear Information System (INIS)

    Tateishi, Ukihide; Maeda, Tetsuo; Morimoto, Tsuyoshi; Miyake, Mototaka; Arai, Yasuaki; Kim, E.E.

    2007-01-01

    The purpose of the present study was to determine the diagnostic accuracy of non-enhanced CT and contrast-enhanced CT in integrated PET/CT studies for preoperative nodal staging of rectal cancer. Retrospective analysis was performed in 53 patients with pathologically proven rectal cancer who had been referred for preoperative staging. All patients underwent integrated PET/CT consisting of non-enhanced and contrast-enhanced CT followed by whole-body fluorine-18-fluorodeoxyglucose ([ 18 F]FDG) PET. Both non-enhanced and contrast-enhanced PET/CT images were evaluated separately by two observers in consensus. The reference standard was histopathologic results. For nodal staging of rectal cancer, we compared diagnostic accuracy on a per-patient basis between the two modalities. Nodal staging was correctly determined with non-enhanced studies in 37 patients (70%) and with contrast-enhanced studies in 42 patients (79%). On a per-patient basis, sensitivity, specificity, positive predictive value, negative predictive value, and accuracy of regional lymph node staging were 85%, 68%, 83%, 72%, and 79%, respectively, with contrast-enhanced studies, and 85%, 42%, 73%, 62%, and 70%, respectively, with non-enhanced studies. The difference in the accuracy of nodal staging between the two modalities was not significant (p = 0.063). Compared with non-enhanced studies, contrast-enhanced studies determined more correctly the status of pararectal lymph nodes (p 0.002), internal iliac lymph nodes (p = 0.004), and obturator lymph nodes (p < 0.0001). Contrast-enhanced PET/CT is superior to non-enhanced PET/CT for precise definition of regional nodal status in rectal cancer. (orig.)

  12. Diagnostic accuracy of the detection of bone change using panoramic TMJ projection. Comparative study with limited cone-beam CT

    International Nuclear Information System (INIS)

    Araki, Kazuyuki; Okano, Tomohiro; Kataoka, Ryuta; Honda, Kazuya; Endo, Atsushi; Kaneko, Norikazu; Funahashi, Itsuo

    2008-01-01

    Panoramic temporoman joint (TMJ) projection is one of the alternative methods of conventional radiography, such as transcranial projection, for diagnosing temporomandibular joint disorder. There have been a few reports describing the diagnostic ability of this method. We evaluated the diagnostic accuracy of detecting bone change with panoramic TMJ projection. Fifty TMJs in 25 patients were examined. All TMJs were examined by panoramic TMJ projection (Hyper XF) and limited cone-beam CT (3D Accuitomo FPD; 3DX). Two observers evaluated the presence of bone change in the TMJ region using panoramic TMJ projection. One other observer evaluated the limited cone-beam CT for the presence and the pattern of bone changes in the TMJ region as the gold standard. Panoramic TMJ findings were evaluated with regard to sensitivity, specificity, and accuracy. Sensitivity, specificity and accuracy of the panoramic TMJ projection were 0.86, 0.76, and 0.82, respectively. These results and those of previous reports on other radiographic methods for TMJ suggest that panoramic TMJ projection is a useful method of screening for bone change due to TMJ disorder. (author)

  13. 68Ga-PSMA PET/CT in Patients with Rising Prostatic-Specific Antigen After Definitive Treatment of Prostate Cancer: Detection Efficacy and Diagnostic accuracy.

    Science.gov (United States)

    Hamed, Maged Abdel Galil; Basha, Mohammad Abd Alkhalik; Ahmed, Hussien; Obaya, Ahmed Ali; Afifi, Amira Hamed Mohamed; Abdelbary, Eman H

    2018-06-20

    68 Ga-prostate-specific membrane antigen-11 ( 68 Ga-PSMA-11) is a recently developed positron emission tomography (PET) tracer that can detect prostate cancer (PC) relapses and metastases with high contrast resolution. The aim of this study was to assess the detection efficacy and diagnostic accuracy of 68 Ga-PSMA PET/CT image in patients with rising prostatic-specific antigen (PSA) after treatment of PC. The present prospective study included 188 patients who exhibited rising of PSA level on a routine follow-up examination after definitive treatment of PC. All patients underwent a 68 Ga-PSMA PET/CT examination. For each patient, we determined the disease stage, the Gleason score, and the maximum standardized uptake value of the local recurrence and extraprostatic metastases. The detection efficacy and diagnostic accuracy of 68 Ga-PSMA PET/CT were established by histopathology and clinical and imaging follow-up as the reference standards. 68 Ga-PSMA PET/CT detected tumour relapse in 165 patients (35 patients had local recurrence, 106 patients had extraprostatic metastases, and 24 patients had combined lesions). The sensitivity, specificity, and accuracy values of 68 Ga-PSMA PET/CT examination in the detection of PC recurrence were 98.8%, 100%, and 98.8%, respectively. 68 Ga-PSMA PET/CT revealed an overall detection rate of 87.8% (165/188) in patients with rising PSA (median of 2.2 ng/mL, and range of 0.01-70 ng/mL). 68 Ga-PSMA PET/CT is a valuable tool for the detection of PC local recurrence or extraprostatic metastases following rising PSA levels after primary definitive therapy and should be incorporated during routine work-up. Copyright © 2018. Published by Elsevier Inc.

  14. Diagnostic accuracy of dual energy CT angiography in patients with diabetes mellitus

    International Nuclear Information System (INIS)

    Schabel, C.; Bongers, M.N.; Syha, R.; Ketelsen, D.; Homann, G.; Notohamiprodjo, M.; Nikolaou, K.; Bamberg, F.; Thomas, C.

    2015-01-01

    Peripheral arterial disease (PAD) represents a major and highly prevalent complication in patients with diabetes mellitus. The diagnostic, non-invasive work-up by computed tomography angiography (CTA) is limited in the presence of extensive calcification. The aim of the study was to determine the diagnostic accuracy of dual energy CTA (DE-CTA) for the detection and characterization of PAD in patients with diabetes mellitus. In this study 30 diabetic patients with suspected or known PAD were retrospectively included in the analysis. All subjects underwent DE-CTA (Somatom Definition Flash, Siemens Healthcare, Erlangen, Germany) prior to invasive angiography, which served as the reference standard. Blinded analysis included assessment of the presence and degree of peripheral stenosis on curved multiplanar reformatting (MPR) and maximum intensity projections (MIP). Conventional measures of diagnostic accuracy were derived. Among the 30 subjects included in the analysis (83 % male, mean age 70.0 ± 10.5 years, 83 % diabetes type 2), the prevalence of critical stenosis in 331 evaluated vessel segments was high (30 %). Dual energy CT identified critical stenoses with a high sensitivity and good specificity using curved MPR (100 % and 93.1 %, respectively) and MIP images (99 % and 91.8 %, respectively). In stratified analysis, the diagnostic accuracy was higher for stenosis pertaining to the pelvic and thigh vessels as compared with the lower extremities (curved MPR accuracy 97.1 % vs. 99.2 vs. 90.9 %; respectively, p < 0.001). The use of DE-CTA allows reliable detection and characterization of peripheral arterial stenosis in patients with diabetes mellitus with higher accuracy in vessels in the pelvic and thigh regions compared with the vessels in the lower legs. (orig.) [de

  15. How far can the radiation dose be lowered in head CT with iterative reconstruction? Analysis of imaging quality and diagnostic accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Wu, Tung-Hsin; Sun, Jing-Yi [National Yang-Ming University, Department of Biomedical Imaging and Radiological Sciences, Taipei (China); Hung, Sheng-Che; Lin, Chung-Jung; Chiu, Chen Fen; Liu, Min-Jsuan; Teng, Michael Mu Huo; Guo, Wan-Yuo; Chang, Cheng-Yen [Taipei Veterans General Hospital, Department of Radiology, Taipei (China); National Yang-Ming University, School of Medicine, Taipei (China); Lin, Chung-Hsien [National Taiwan University, Graduate Institute of Epidemiology and Preventive Medicine, Taipei (China)

    2013-09-15

    To evaluate the imaging quality of head CT at lowered radiation dose by combining filtered back projection (FBP) and iterative reconstruction (IR) algorithms. Experimental group A (n = 66) underwent CT with 43 % tube current reduction, and group B (n = 58) received an equivalent reduced dose by lowering the tube voltage. An age- and sex-matched control group (n = 72) receiving the conventional radiation dose was retrospectively collected. Imaging for the control group was reconstructed by FBP only, while images for groups A and B were reconstructed by FBP and IR. The signal-to-noise ratios (SNRs), contrast-to-noise ratios (CNRs), sharpness, number of infarcts and severity of subcortical arteriosclerotic encephalopathy (SAE) were compared to assess imaging quality and diagnostic accuracy. There were no significant differences in SNRs and CNRs between group A and the control group. There were significantly decreased SNRs and increased CNRs in group B. Image sharpness decreased in both groups. Correlations between detected infarcts and severity of SAE across FBP and IR were high (r = 0.73-0.93). Head diameter was the only significant factor inversely correlated with infratentorial imaging quality. Head CT with 43 % reduced tube current reconstructed by IR provides diagnostic imaging quality for outpatient management. (orig.)

  16. In-vitro study on the accuracy of a simple-design CT-guided stent for dental implants

    International Nuclear Information System (INIS)

    Huh, Young June; Choi, Bo Ram; Huh, Kyung Hoe; Yi, Won Jin; Heo, Min Suk; Lee, Sam Sun; Choi, Soon Chul

    2012-01-01

    An individual surgical stent fabricated from computed tomography (CT) data, called a CT-guided stent, would be useful for accurate installation of implants. The purpose of the present study was to introduce a newly developed CT-guided stent with a simple design and evaluate the accuracy of the stent placement. A resin template was fabricated from a hog mandible and a specially designed plastic plate, with 4 metal balls inserted in it for radiographic recognition, was attached to the occlusal surface of the template. With the surgical stent applied, CT images were taken, and virtual implants were placed using software. The spatial positions of the virtually positioned implants were acquired and implant guiding holes were drilled into the surgical stent using a specially designed 5-axis drilling machine. The surgical stent was placed on the mandible and CT images were taken again. The discrepancy between the central axis of the drilled holes on the second CT images and the virtually installed implants on the first CT images was evaluated. The deviation of the entry point and angulation of the central axis in the reference plane were 0.47±0.27 mm, 0.57±0.23 mm, and 0.64±0.16 degree, 0.57±0.15 degree, respectively. However, for the two different angulations in each group, the 20 degree angulation showed a greater error in the deviation of the entry point than did the 10 degree angulation. The CT-guided template proposed in this study was highly accurate. It could replace existing implant guide systems to reduce costs and effort.

  17. Diagnostic accuracy of low-dose CT compared with abdominal radiography in non-traumatic acute abdominal pain: prospective study and systematic review

    International Nuclear Information System (INIS)

    Alshamari, Muhammed; Geijer, Haakan; Norrman, Eva; Geijer, Mats; Jansson, Kjell

    2016-01-01

    Abdominal radiography is frequently used in acute abdominal non-traumatic pain despite the availability of more advanced diagnostic modalities. This study evaluates the diagnostic accuracy of low-dose CT compared with abdominal radiography, at similar radiation dose levels. Fifty-eight patients were imaged with both methods and were reviewed independently by three radiologists. The reference standard was obtained from the diagnosis in medical records. Sensitivity and specificity were calculated. A systematic review was performed after a literature search, finding a total of six relevant studies including the present. Overall sensitivity with 95 % CI for CT was 75 % (66-83 %) and 46 % (37-56 %) for radiography. Specificity was 87 % (77-94 %) for both methods. In the systematic review the overall sensitivity for CT varied between 75 and 96 % with specificity from 83 to 95 % while the overall sensitivity for abdominal radiography varied between 30 and 77 % with specificity 75 to 88 %. Based on the current study and available evidence, low-dose CT has higher diagnostic accuracy than abdominal radiography and it should, where logistically possible, replace abdominal radiography in the workup of adult patients with acute non-traumatic abdominal pain. (orig.)

  18. Diagnostic accuracy of low-dose CT compared with abdominal radiography in non-traumatic acute abdominal pain: prospective study and systematic review

    Energy Technology Data Exchange (ETDEWEB)

    Alshamari, Muhammed; Geijer, Haakan [Oerebro University, Department of Radiology, Faculty of Medicine and Health, Oerebro (Sweden); Norrman, Eva [Oerebro University, Department of Medical Physics, Faculty of Medicine and Health, Oerebro (Sweden); Geijer, Mats [Lund University and Skaane University Hospital, Department of Medical Imaging and Physiology, Lund (Sweden); Jansson, Kjell [Oerebro University, Department of Surgery, Faculty of Medicine and Health, Oerebro (Sweden)

    2016-06-15

    Abdominal radiography is frequently used in acute abdominal non-traumatic pain despite the availability of more advanced diagnostic modalities. This study evaluates the diagnostic accuracy of low-dose CT compared with abdominal radiography, at similar radiation dose levels. Fifty-eight patients were imaged with both methods and were reviewed independently by three radiologists. The reference standard was obtained from the diagnosis in medical records. Sensitivity and specificity were calculated. A systematic review was performed after a literature search, finding a total of six relevant studies including the present. Overall sensitivity with 95 % CI for CT was 75 % (66-83 %) and 46 % (37-56 %) for radiography. Specificity was 87 % (77-94 %) for both methods. In the systematic review the overall sensitivity for CT varied between 75 and 96 % with specificity from 83 to 95 % while the overall sensitivity for abdominal radiography varied between 30 and 77 % with specificity 75 to 88 %. Based on the current study and available evidence, low-dose CT has higher diagnostic accuracy than abdominal radiography and it should, where logistically possible, replace abdominal radiography in the workup of adult patients with acute non-traumatic abdominal pain. (orig.)

  19. TH-CD-202-04: Evaluation of Virtual Non-Contrast Images From a Novel Split-Filter Dual-Energy CT Technique

    International Nuclear Information System (INIS)

    Huang, J; Szczykutowicz, T; Bayouth, J; Miller, J

    2016-01-01

    Purpose: To compare the ability of two dual-energy CT techniques, a novel split-filter single-source technique of superior temporal resolution against an established sequential-scan technique, to remove iodine contrast from images with minimal impact on CT number accuracy. Methods: A phantom containing 8 tissue substitute materials and vials of varying iodine concentrations (1.7–20.1 mg I /mL) was imaged using a Siemens Edge CT scanner. Dual-energy virtual non-contrast (VNC) images were generated using the novel split-filter technique, in which a 120kVp spectrum is filtered by tin and gold to create high- and low-energy spectra with < 1 second temporal separation between the acquisition of low- and high-energy data. Additionally, VNC images were generated with the sequential-scan technique (80 and 140kVp) for comparison. CT number accuracy was evaluated for all materials at 15, 25, and 35mGy CTDIvol. Results: The spectral separation was greater for the sequential-scan technique than the split-filter technique with dual-energy ratios of 2.18 and 1.26, respectively. Both techniques successfully removed iodine contrast, resulting in mean CT numbers within 60HU of 0HU (split-filter) and 40HU of 0HU (sequential-scan) for all iodine concentrations. Additionally, for iodine vials of varying diameter (2–20 mm) with the same concentration (9.9 mg I /mL), the system accurately detected iodine for all sizes investigated. Both dual-energy techniques resulted in reduced CT numbers for bone materials (by >400HU for the densest bone). Increasing the imaging dose did not improve the CT number accuracy for bone in VNC images. Conclusion: VNC images from the split-filter technique successfully removed iodine contrast. These results demonstrate a potential for improving dose calculation accuracy and reducing patient imaging dose, while achieving superior temporal resolution in comparison sequential scans. For both techniques, inaccuracies in CT numbers for bone materials

  20. TH-CD-202-04: Evaluation of Virtual Non-Contrast Images From a Novel Split-Filter Dual-Energy CT Technique

    Energy Technology Data Exchange (ETDEWEB)

    Huang, J; Szczykutowicz, T; Bayouth, J; Miller, J [University of Wisconsin, Madison, WI (United States)

    2016-06-15

    Purpose: To compare the ability of two dual-energy CT techniques, a novel split-filter single-source technique of superior temporal resolution against an established sequential-scan technique, to remove iodine contrast from images with minimal impact on CT number accuracy. Methods: A phantom containing 8 tissue substitute materials and vials of varying iodine concentrations (1.7–20.1 mg I /mL) was imaged using a Siemens Edge CT scanner. Dual-energy virtual non-contrast (VNC) images were generated using the novel split-filter technique, in which a 120kVp spectrum is filtered by tin and gold to create high- and low-energy spectra with < 1 second temporal separation between the acquisition of low- and high-energy data. Additionally, VNC images were generated with the sequential-scan technique (80 and 140kVp) for comparison. CT number accuracy was evaluated for all materials at 15, 25, and 35mGy CTDIvol. Results: The spectral separation was greater for the sequential-scan technique than the split-filter technique with dual-energy ratios of 2.18 and 1.26, respectively. Both techniques successfully removed iodine contrast, resulting in mean CT numbers within 60HU of 0HU (split-filter) and 40HU of 0HU (sequential-scan) for all iodine concentrations. Additionally, for iodine vials of varying diameter (2–20 mm) with the same concentration (9.9 mg I /mL), the system accurately detected iodine for all sizes investigated. Both dual-energy techniques resulted in reduced CT numbers for bone materials (by >400HU for the densest bone). Increasing the imaging dose did not improve the CT number accuracy for bone in VNC images. Conclusion: VNC images from the split-filter technique successfully removed iodine contrast. These results demonstrate a potential for improving dose calculation accuracy and reducing patient imaging dose, while achieving superior temporal resolution in comparison sequential scans. For both techniques, inaccuracies in CT numbers for bone materials

  1. Diagnostic accuracy of C-arm CT during selective transcatheter angiography for hepatocellular carcinoma: comparison with intravenous contrast-enhanced, biphasic, dynamic MDCT

    Energy Technology Data Exchange (ETDEWEB)

    Higashihara, Hiroki; Osuga, Keigo; Onishi, Hiromitsu; Nakamoto, Atsushi; Tsuboyama, Takahiro; Maeda, Noboru; Hori, Masatoshi; Kim, Tonsok; Tomiyama, Noriyuki [Osaka University Graduate School of Medicine, Department of Diagnostic and Interventional Radiology, Suita, Osaka (Japan)

    2012-04-15

    This study was aimed to compare the accuracy, sensitivity, and positive predictive value of C-arm CT (CACT) during selective transcatheter angiography with those of multidetector CT (MDCT) in the detection of hepatocellular carcinoma (HCC). In this prospective study, 30 patients (mean age, 73 years) with unresectable HCC were examined with CACT before chemoembolisation. Images of a combination of CACT during arterial portography (CACTAP) and dual-phase CACT during hepatic arteriography (CACTHA) was obtained and images of intravenous contrast-enhanced, biphasic, dynamic, MDCT was also obtained beforehand. Three blinded observers independently reviewed CACT and MDCT. Diagnostic accuracy was evaluated by the alternative free-response receiver operating characteristic (AFROC) method. Sensitivities and positive predictive values (PPV) were analyzed with the paired t-test. In the mean area under the AFROC curve (Az), there was no significant difference between MDCT and CACT (MDCT, mean Az value, 0.83; CACT, 0.85, respectively) (P = 0.32). There was also no significant difference between the two techniques in sensitivity (MDCT, mean 0.65; CACT, 0.60) and PPV (MDCT, mean 0.98; CACT, 0.97) (P = 0.40, P = 0.68, respectively). The diagnostic accuracy of CACT was equivalent to that of biphasic CT in the diagnosis of HCC. (orig.)

  2. Diagnostic accuracy of virtual non-contrast enhanced dual-energy CT for diagnosis of adrenal adenoma: A systematic review and meta-analysis

    Energy Technology Data Exchange (ETDEWEB)

    Connolly, Michael J.; McInnes, Matthew D.F.; Schieda, Nicola [University of Ottawa Department of Radiology, Clinical Epidemiology Program, Ottawa Hospital Research Institute, Ottawa, ON (United States); El-Khodary, Mohamed [McMaster University Department of Radiology, Hamilton, ON (Canada); McGrath, Trevor A. [University of Ottawa, Faculty of Medicine, Ottawa, Ontario (Canada)

    2017-10-15

    To compare the diagnostic accuracy of dual-energy (DE) virtual non-contrast computed tomography (vNCT) to non-contrast CT (NCT) for the diagnosis of adrenal adenomas. Search of multiple databases and grey literature was performed. Two reviewers independently applied inclusion criteria and extracted data. Risk of bias was assessed using QUADAS-2. Summary estimates of diagnostic accuracy were generated and sources of heterogeneity were assessed. Five studies (170 patients; 192 adrenal masses) were included for diagnostic accuracy assessment; all used dual-source dual-energy CT. Pooled sensitivity for adrenal adenoma on vNCT was 54% (95% CI: 47-62%). Pooled sensitivity for NCT was 57% (95% CI: 45-69%). Pooling of specificity was not performed since no false positives were reported. There was a trend for overestimation of HU density on vNCT as compared to NCT which appeared related to contrast timing. Potential sources of bias were seen regarding the index test and reference standard for the included studies. Potential sources of heterogeneity between studies were seen in adenoma prevalence and intravenous contrast timing. vNCT images generated from dual-energy CT demonstrated comparable sensitivity to NCT for the diagnosis of adenomas; however the included studies are heterogeneous and at high risk for some types of bias. (orig.)

  3. Assessment of US score and CT number for diagnosis of fatty liver

    International Nuclear Information System (INIS)

    Ogasawara, Tetsuo; Tanda, Shigeru; Lim, Insu; Oota, Keisuke; Taima, Tadashi

    1987-01-01

    The author evaluates US and CT for diagnosis of fatty liver in 70 cases with fatty change of the liver. We tried to score the US findings of the fatty change, i.e., ''bright liver pattern'', ''liver-kidney contrast'', ''vascular blurring'', ''deep attenuation'', and the usefulness of the scoring was examined. Comparing with CT number, US score was more sensitive, but had no significant correlation with the amount of the fat in the liver and with the abnormality of the liver function tests. The results indicate that US should be used as a primary screening examination, and for the further evaluation of the fatty change of the liver, CT should be carried out. (author)

  4. Comparison of Accuracy of Contrast Enhanced Computed Tomography with Accuracy of Non-Contrast Magnetic Resonance Imaging in Evaluation of Local Extension of Base of Tongue Malignancies

    Directory of Open Access Journals (Sweden)

    Ketan Rathod

    2018-01-01

    Full Text Available Diagnosis of base of tongue malignancy can be obtained through clinical examination and biopsy. Magnetic Resonance Imaging (MRI and Computed Tomography (CT are used to detect its local extension, nodal spread and distant metastases. The main aim of study was to compare the accuracy of MRI and contrast enhanced CT in determining the local extent of base of tongue malignancy. Twenty five patients, biopsy proven cases of squamous cell carcinoma of base of tongue were taken. 1.5 Tesla Magnetic Resonance Unit with T2 weighted axial, coronal image; T1 weighted axial, coronal image; and STIR (Short tau inversion recovery axial and coronal images were used. 16 slice Computed Tomography unit with non-contrast and contrast enhanced images were used. Accuracy of CT to detect midline crossing: 50%; accuracy of MRI to detect midline crossing: 100%; accuracy of CT to detect anterior extension: 92%; accuracy of MRI to detect anterior extension: 100%; accuracy of CT to detect tonsillar fossa invasion: 83%; accuracy of MRI to detect tonsillar fossa invasion: 100%; accuracy of CT to detect oro pharyngeal spread: 83%; accuracy of MRI to detect oro pharyngeal spread: 100%; accuracy of CT to detect bone involvement: 20%; accuracy of MRI to detect bone involvement: 100%. MRI proved to be a better investigation than CT, in terms of evaluation of depth of invasion, presence of bony involvement, extension to opposite side, anterior half of tongue, tonsillar fossa, floor of mouth or oropharynx.

  5. CT quantitative diagnosis in fatty liver: a clinical study

    International Nuclear Information System (INIS)

    He Wen; Qian Linxue; Zhao Jixue; Ma Daqing; Feng Jie; Hu Zhihai

    2001-01-01

    Objective: To establish the CT criteria of quantitative diagnosis for liver steatosis by means of studying the CT features of fatty liver cases proven histologically. Methods: Twenty-eight cases of fatty liver were underwent non-enhanced CT scan, and the attenuation of liver parenchyma was measured. To differentiate the degree of fatty liver, the mean CT value and the relative density of hepatic vessels were observed. The quantitative diagnosis was made according to the CT number threshold and the criteria of relative density of hepatic vessels, respectively. Results: Among the 28 cases, there were 17 cases of mild steatosis with mean CT number of 46 HU (32-65 HU), 7 cases of middle degree fatty liver with mean CT number of 28 HU (15-38 HU), and 4 cases of sever fatty liver with mean CT number of 0.2 HU (-7-11 HU). For the relative density of hepatic vessels, 16 of the 17 cases of mild fatty liver had a appearance of hepatic vessels immersion and 1 mild case had reverse hepatic vessels display, 6 of 7 middle degree cases had reverse hepatic vessels display with 1 case having the appearance of hepatic vessels immersion, and all the 4 case of sever steatosis had the appearance of reverse hepatic vessels display with sharp contrast between vessels and the liver parenchyma. The accuracy of quantitative diagnosis was 65.9% and 93.1% by means of criteria of CT number threshold and relative density of hepatic vessels, respectively (x 2 = 7.153, P < 0.01). Conclusion: The criteria of relative density of hepatic vessels is more reliable than that of CT number threshold in quantitative diagnosis of fatty liver

  6. Automatic radiation dose monitoring for CT of trauma patients with different protocols: feasibility and accuracy

    International Nuclear Information System (INIS)

    Higashigaito, K.; Becker, A.S.; Sprengel, K.; Simmen, H.-P.; Wanner, G.; Alkadhi, H.

    2016-01-01

    Aim: To demonstrate the feasibility and accuracy of automatic radiation dose monitoring software for computed tomography (CT) of trauma patients in a clinical setting over time, and to evaluate the potential of radiation dose reduction using iterative reconstruction (IR). Materials and methods: In a time period of 18 months, data from 378 consecutive thoraco-abdominal CT examinations of trauma patients were extracted using automatic radiation dose monitoring software, and patients were split into three cohorts: cohort 1, 64-section CT with filtered back projection, 200 mAs tube current–time product; cohort 2, 128-section CT with IR and identical imaging protocol; cohort 3, 128-section CT with IR, 150 mAs tube current–time product. Radiation dose parameters from the software were compared with the individual patient protocols. Image noise was measured and image quality was semi-quantitatively determined. Results: Automatic extraction of radiation dose metrics was feasible and accurate in all (100%) patients. All CT examinations were of diagnostic quality. There were no differences between cohorts 1 and 2 regarding volume CT dose index (CTDI_v_o_l; p=0.62), dose–length product (DLP), and effective dose (ED, both p=0.95), while noise was significantly lower (chest and abdomen, both −38%, p<0.017). Compared to cohort 1, CTDI_v_o_l, DLP, and ED in cohort 3 were significantly lower (all −25%, p<0.017), similar to the noise in the chest (–32%) and abdomen (–27%, both p<0.017). Compared to cohort 2, CTDI_v_o_l (–28%), DLP, and ED (both –26%) in cohort 3 was significantly lower (all, p<0.017), while noise in the chest (+9%) and abdomen (+18%) was significantly higher (all, p<0.017). Conclusion: Automatic radiation dose monitoring software is feasible and accurate, and can be implemented in a clinical setting for evaluating the effects of lowering radiation doses of CT protocols over time. - Highlights: • Automatic dose monitoring software can be

  7. SPECT/CT in pediatric patient management

    International Nuclear Information System (INIS)

    Nadel, Helen R.

    2014-01-01

    Hybrid SPECT/CT imaging is becoming the standard of care in pediatric imaging. Indications are mainly for oncologic imaging including mIBG scintigraphy for neuroblastoma and I-123 post surgical imaging of children with thyroid carcinoma, bone scintigraphy for back pain, children referred from sports medicine and neurodevelopmentally delayed children presenting with pain symptoms. The studies provide improved diagnostic accuracy, and oncologic imaging that includes optimized CT as part of the SPECT/CT study may decrease the number of studies and sedation procedures an individual child may need. The studies, however, must be tailored on an individual basis as the addition of the CT study can increase exposure to the child and should only be performed after appropriate justification and with adherence to optimized low dose pediatric protocols. (orig.)

  8. SPECT/CT in pediatric patient management

    Energy Technology Data Exchange (ETDEWEB)

    Nadel, Helen R. [British Columbia Children' s Hospital, University of British Columbia, Pediatric Radiologist and Nuclear Medicine Physician, Division of Nuclear Medicine Department of Radiology, Vancouver, British Columbia (Canada)

    2014-05-15

    Hybrid SPECT/CT imaging is becoming the standard of care in pediatric imaging. Indications are mainly for oncologic imaging including mIBG scintigraphy for neuroblastoma and I-123 post surgical imaging of children with thyroid carcinoma, bone scintigraphy for back pain, children referred from sports medicine and neurodevelopmentally delayed children presenting with pain symptoms. The studies provide improved diagnostic accuracy, and oncologic imaging that includes optimized CT as part of the SPECT/CT study may decrease the number of studies and sedation procedures an individual child may need. The studies, however, must be tailored on an individual basis as the addition of the CT study can increase exposure to the child and should only be performed after appropriate justification and with adherence to optimized low dose pediatric protocols. (orig.)

  9. Image quality, radiation dose and diagnostic accuracy of 70 kVp whole brain volumetric CT perfusion imaging: a preliminary study

    Energy Technology Data Exchange (ETDEWEB)

    Fang, Xiao Kun; Ni, Qian Qian; Zhou, Chang Sheng; Chen, Guo Zhong; Luo, Song; Zhang, Long Jiang; Lu, Guang Ming [Medical School of Nanjing University, Department of Medical Imaging, Jinling Hospital, Nanjing, Jiangsu (China); Schoepf, U.J. [Medical School of Nanjing University, Department of Medical Imaging, Jinling Hospital, Nanjing, Jiangsu (China); Medical University of South Carolina, Ashley River Tower, Division of Cardiovascular Imaging, Charleston, SC (United States); Fuller, Stephen R.; De Cecco, Carlo N. [Medical University of South Carolina, Ashley River Tower, Division of Cardiovascular Imaging, Charleston, SC (United States)

    2016-11-15

    To evaluate image quality and diagnostic accuracy for acute infarct detection and radiation dose of 70 kVp whole brain CT perfusion (CTP) and CT angiography (CTA) reconstructed from CTP source data. Patients were divided into three groups (n = 50 each): group A, 80 kVp, 21 scanning time points; groups B, 70 kVp, 21 scanning time points; group C, 70 kVp, 17 scanning time points. Objective and subjective image quality of CTP and CTA were compared. Diagnostic accuracy for detecting acute infarct and cerebral artery stenosis ≥ 50 % was calculated for CTP and CTA with diffusion weighted imaging and digital subtraction angiography as reference standards. Effective radiation dose was compared. There were no differences in any perfusion parameter value between three groups (P > 0.05). No difference was found in subjective image quality between three groups (P > 0.05). Diagnostic accuracy for detecting acute infarct and vascular stenosis showed no difference between three groups (P > 0.05). Compared with group A, radiation doses of groups B and C were decreased by 28 % and 37 % (both P < 0.001), respectively. Compared with 80 kVp protocol, 70 kVp brain CTP allows comparable vascular and perfusion assessment and lower radiation dose while maintaining high diagnostic accuracy in detecting acute infarct. (orig.)

  10. Diagnostic Accuracy of Ultrasound, Contrast-enhanced CT, and Conventional MRI for Differentiating Leiomyoma From Leiomyosarcoma.

    Science.gov (United States)

    Gaetke-Udager, Kara; McLean, Karen; Sciallis, Andrew P; Alves, Timothy; Maturen, Katherine E; Mervak, Benjamin M; Moore, Andreea G; Wasnik, Ashish P; Erba, Jake; Davenport, Matthew S

    2016-10-01

    This study aimed to determine whether uterine leiomyoma can be distinguished from uterine leiomyosarcoma on ultrasound (US), computed tomography (CT), and/or magnetic resonance imaging (MRI) without diffusion-weighted imaging. Institutional review board approval was obtained and informed consent was waived for this Health Insurance Portability and Accountability Act-compliant retrospective case-control diagnostic accuracy study. All subjects with resected uterine leiomyosarcoma diagnosed over a 17-year period (1998-2014) at a single institution for whom pre-resection US (n = 10), CT (n = 11), or MRI (n = 7) was available were matched by tumor size and imaging modality with 28 subjects with resected uterine leiomyoma. Six blinded radiologists (three attendings, three residents) assigned 5-point Likert scores for the following features: (1) margins, (2) necrosis, (3) hemorrhage, (4) vascularity, (5) calcifications, (6) heterogeneity, and (7) likelihood of malignancy (primary end point). Mean suspicion scores were calculated and receiver operating characteristic curves were generated. The ability of individual morphologic features to predict malignancy was assessed with logistic regression. Mean suspicion scores were 2.5 ± 1.2 (attendings) and 2.4 ± 1.3 (residents) for leiomyoma, and 2.7 ± 1.3 (attendings) and 2.7 ± 1.4 (residents) for leiomyosarcoma. The areas under the receiver operating characteristic curves (range: 0.330-0.685) were not significantly different from chance, either overall (P = .36-.88) or by any modality (P = .28-.96), for any reader. Reader experience had no effect on diagnostic accuracy. No morphologic parameter was significantly predictive of malignancy (P = .10-.97). Uterine leiomyoma cannot be differentiated accurately from leiomyosarcoma on US, CT, or MRI without diffusion-weighted imaging. Copyright © 2016 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

  11. Diagnostic accuracy of coronary CT angiography using 3{sup rd}-generation dual-source CT and automated tube voltage selection: Clinical application in a non-obese and obese patient population

    Energy Technology Data Exchange (ETDEWEB)

    Mangold, Stefanie [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); Eberhard-Karls University Tuebingen, Department of Diagnostic and Interventional Radiology, Tuebingen (Germany); Wichmann, Julian L. [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); University Hospital Frankfurt, Department of Diagnostic and Interventional Radiology, Frankfurt (Germany); Schoepf, U.J. [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); Medical University of South Carolina, Division of Cardiology, Department of Medicine, Charleston, SC (United States); Caruso, Damiano [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); University of Rome ' ' Sapienza' ' , Department of Radiological Sciences, Oncology and Pathology, Rome (Italy); Tesche, Christian [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); Heart Centre Munich-Bogenhausen, Department of Cardiology, Munich (Germany); Steinberg, Daniel H.; Bayer, Richard R. [Medical University of South Carolina, Division of Cardiology, Department of Medicine, Charleston, SC (United States); Varga-Szemes, Akos; Stubenrauch, Andrew C.; Biancalana, Matthew; De Cecco, Carlo N. [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); Nikolaou, Konstantin [Eberhard-Karls University Tuebingen, Department of Diagnostic and Interventional Radiology, Tuebingen (Germany)

    2017-06-15

    To investigate diagnostic accuracy of 3{sup rd}-generation dual-source CT (DSCT) coronary angiography in obese and non-obese patients. We retrospectively analyzed 76 patients who underwent coronary CT angiography (CCTA) and invasive coronary angiography. Prospectively ECG-triggered acquisition was performed with automated tube voltage selection (ATVS). Patients were dichotomized based on body mass index in groups A (<30 kg/m{sup 2}, n = 37) and B (≥30 kg/m{sup 2}, n = 39) and based on tube voltage in groups C (<120 kV, n = 46) and D (120 kV, n = 30). Coronary arteries were assessed for significant stenoses (≥50 % luminal narrowing) and diagnostic accuracy was calculated. Per-patient overall sensitivity, specificity, positive predictive value, negative predictive value (NPV) and accuracy were 96.9 %, 95.5 %, 93.9 %, 97.7 % and 96.1 %, respectively. Sensitivity and NPV were lower in groups B and D compared to groups A and C, but no statistically significant differences were observed (group A vs. B: sensitivity, 100.0 % vs. 93.3 %, p = 0.9493; NPV, 100 % vs. 95.5 %, p = 0.9812; group C vs. D: sensitivity, 100.0 % vs. 92.3 %, p = 0.8462; NPV, 100.0 % vs. 94.1 %, p = 0.8285). CCTA using 3{sup rd}-generation DSCT and (ATVS) provides high diagnostic accuracy in both non-obese and obese patients. (orig.)

  12. Diagnostic accuracy of coronary CT angiography using 3"r"d-generation dual-source CT and automated tube voltage selection: Clinical application in a non-obese and obese patient population

    International Nuclear Information System (INIS)

    Mangold, Stefanie; Wichmann, Julian L.; Schoepf, U.J.; Caruso, Damiano; Tesche, Christian; Steinberg, Daniel H.; Bayer, Richard R.; Varga-Szemes, Akos; Stubenrauch, Andrew C.; Biancalana, Matthew; De Cecco, Carlo N.; Nikolaou, Konstantin

    2017-01-01

    To investigate diagnostic accuracy of 3"r"d-generation dual-source CT (DSCT) coronary angiography in obese and non-obese patients. We retrospectively analyzed 76 patients who underwent coronary CT angiography (CCTA) and invasive coronary angiography. Prospectively ECG-triggered acquisition was performed with automated tube voltage selection (ATVS). Patients were dichotomized based on body mass index in groups A (<30 kg/m"2, n = 37) and B (≥30 kg/m"2, n = 39) and based on tube voltage in groups C (<120 kV, n = 46) and D (120 kV, n = 30). Coronary arteries were assessed for significant stenoses (≥50 % luminal narrowing) and diagnostic accuracy was calculated. Per-patient overall sensitivity, specificity, positive predictive value, negative predictive value (NPV) and accuracy were 96.9 %, 95.5 %, 93.9 %, 97.7 % and 96.1 %, respectively. Sensitivity and NPV were lower in groups B and D compared to groups A and C, but no statistically significant differences were observed (group A vs. B: sensitivity, 100.0 % vs. 93.3 %, p = 0.9493; NPV, 100 % vs. 95.5 %, p = 0.9812; group C vs. D: sensitivity, 100.0 % vs. 92.3 %, p = 0.8462; NPV, 100.0 % vs. 94.1 %, p = 0.8285). CCTA using 3"r"d-generation DSCT and (ATVS) provides high diagnostic accuracy in both non-obese and obese patients. (orig.)

  13. Evaluation of CT in hepatocellular carcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Nakao, Norio; Miura, Yukio; Ohnishi, Mitsunori; Kamikon-ya, Norihiko; Sakamoto, Yoshisato; Miura, Takashi; Sakamoto, Kiyoshi; Takayasu, Yoshio

    1985-06-01

    In order to evaluate the diagnostic ability of CT in hepatocellular carcinoma, four kinds of CT images were comparatively studied by the accuracy and ROC (receiver operating characteristic) curve analysis. As a result, it was clarified that CT images were evaluated more objectively by ROC curve analysis than by accuracy. The diagnostic ability of existence and differentiation of tumor in the liver were higher in order of plain CT, contrast enhanced CT (CECT), bolus CT and CT arteriography (CTA). Therefore, in an usual CT examination intended to make differential diagnosis in space occupying liver disease, bolus CT seems to be indispensable, and also CTA is essential where diagnosis is difficult even by bolus CT.

  14. Effects of x-ray and CT image enhancements on the robustness and accuracy of a rigid 3D/2D image registration

    International Nuclear Information System (INIS)

    Kim, Jinkoo; Yin Fangfang; Zhao Yang; Kim, Jae Ho

    2005-01-01

    A rigid body three-dimensional/two-dimensional (3D/2D) registration method has been implemented using mutual information, gradient ascent, and 3D texturemap-based digitally reconstructed radiographs. Nine combinations of commonly used x-ray and computed tomography (CT) image enhancement methods, including window leveling, histogram equalization, and adaptive histogram equalization, were examined to assess their effects on accuracy and robustness of the registration method. From a set of experiments using an anthropomorphic chest phantom, we were able to draw several conclusions. First, the CT and x-ray preprocessing combination with the widest attraction range was the one that linearly stretched the histograms onto the entire display range on both CT and x-ray images. The average attraction ranges of this combination were 71.3 mm and 61.3 deg in the translation and rotation dimensions, respectively, and the average errors were 0.12 deg and 0.47 mm. Second, the combination of the CT image with tissue and bone information and the x-ray images with adaptive histogram equalization also showed subvoxel accuracy, especially the best in the translation dimensions. However, its attraction ranges were the smallest among the examined combinations (on average 36 mm and 19 deg). Last the bone-only information on the CT image did not show convergency property to the correct registration

  15. Software-based hybrid perfusion SPECT/CT provides diagnostic accuracy when other pulmonary embolism imaging is indeterminate

    Energy Technology Data Exchange (ETDEWEB)

    Kumar, Nishant; Xie, Karen; Mar, Winnie; Anderson, Thomas M.; Carney, Benjamin; Mehta, Nikhil; Machado, Roberto; Blend, Michael J.; Lu, Yang [University of Illinois Hospital and Health Sciences System, Chicago (Korea, Republic of)

    2015-12-15

    To investigate the diagnostic performance of perfusion single-photon emission computed tomography/computed tomography (Q-SPECT/CT) in patients suspected to have pulmonary embolism (PE) but with indeterminate computed tomographic pulmonary angiography (CTPA) or planar ventilation/perfusion (V/Q) scans. This retrospective study included two groups of patients. Group I consisted of 49 patients with nondiagnostic CTPA. These 49 patients underwent subsequent V/Q scans. Further Q-SPECTs were obtained in patients with indeterminate planar images and fused with existing CTPA. Group II consisted of 182 non-CTPA patients with indeterminate V/Q scans. These 182 patients underwent further Q-SPECT and separate noncontrast low-dose CT chest. Fusion Q-SPECT/CT scans were obtained through FDA-approved software and interpreted according to published criteria as positive, negative, or indeterminate for PE. Upon retrospective analyses, the final diagnosis was made using composite reference standards including all available clinical and imaging information for at least 6-month follow-up. In group I patients, 1 was positive, 24 were negative, and another 24 (49 %, 24/49) were indeterminate. In the subsequent 24 Q-SPECT/CTPAs, 4 were positive, 19 were negative, and 1 was indeterminate (4.2 %, 1/24). In group II patients, 9 (4.9 %, 9/182) were indeterminate, 33 were positive, and 140 were negative. The combined nondiagnostic rate for Q-SPECT/CT was only 4.9 % (10/206). There was six false-negative and one false-positive Q-SPECT/CT examinations. The sensitivity, specificity, and positive and negative predictive value of Q-SPECT/CT were 85.7 % (36/42), 99.4 % (153/154), 97.3 % (36/37) and 96.2 % (153/159), respectively. Q-SPECT/CT improves the diagnostic rate with promising accuracy in diagnosing PE that yields a satisfactory clinical verdict, especially when the CTPA and planar V/Q scan are indeterminate.

  16. Accuracy and reliability of thyroid volumetry using spiral CT and thyroid volume in a healthy, non-iodine-deficient Chinese adult population

    Energy Technology Data Exchange (ETDEWEB)

    Shu Jian, E-mail: shujiannc@163.com [Department of Radiology, Second Affiliated Hospital of Chongqing Medical University, 74 Linjiang Road, Yuzhong District, Chongqing 400010 (China); Zhao Jiannong, E-mail: zhaojiannong@tom.com [Department of Radiology, Second Affiliated Hospital of Chongqing Medical University, 74 Linjiang Road, Yuzhong District, Chongqing 400010 (China); Guo Dajing, E-mail: guodaj@163.com [Department of Radiology, Second Affiliated Hospital of Chongqing Medical University, 74 Linjiang Road, Yuzhong District, Chongqing 400010 (China); Luo Yindeng, E-mail: yindengluo_1019@163.com [Department of Radiology, Second Affiliated Hospital of Chongqing Medical University, 74 Linjiang Road, Yuzhong District, Chongqing 400010 (China); Zhong Weijia, E-mail: zhongweijia2003@eyou.com [Department of Radiology, Second Affiliated Hospital of Chongqing Medical University, 74 Linjiang Road, Yuzhong District, Chongqing 400010 (China); Xie Weibo, E-mail: radiologycq@163.com [Department of Radiology, Second Affiliated Hospital of Chongqing Medical University, 74 Linjiang Road, Yuzhong District, Chongqing 400010 (China)

    2011-02-15

    Objective: The purpose of this study was to assess the accuracy and reliability of thyroid volumetry using spiral CT and to investigate thyroid volumes for a healthy, non-iodine-deficient adult population in southwestern region of China. Materials and methods: Spiral CT was performed in phantoms and adult subjects with normal thyroid, and the volumes were measured by observers with 5 years or more of CT experience. The phantom volumes and the thyroid volumes of all subjects were noted. Results: For the thyroid phantoms, there was no significant difference between the true and CT calculated volumes (t = 0.862, P = 0.399), and the correlation was excellent (ICC = 0.9995, P = 0.000). In the subjects for reliability analysis, the intraobserver or interobserver differences for CT volumetric measurement of thyroid were not significant (P > 0.05). The intraobserver or interobserver correlations were very high (ICC > 0.99, P < 0.001). In the subjects for population analysis, the median of the thyroid volumes was 11.45 cm{sup 3}. The nonparametric Mann-Whitney U-test showed no significant difference for the thyroid volume between sexes (U = 4388.00, Z = -1.118, P = 0.264). The nonparametric Kruskall-Wallis test showed no significant difference in all age groups ({chi}{sup 2} = 13.466, P = 0.062). There was a slight negative correlation between the thyroid volume and age (r{sub s} = -0.166, P = 0.019). Conclusion: The accuracy and reliability of multi-slice spiral CT in measuring thyroid volume are very high. The thyroid volumes are not significantly difference between genders or among decades for the healthy, non-iodine-deficient adult population in southwestern region of China.

  17. Detection of pulmonary metastases with pathological correlation: effect of breathing on the accuracy of spiral CT. Editor's note

    International Nuclear Information System (INIS)

    Coakley, F.V.; Cohen, M.D.; Waters, D.J.; Davis, M.M.; Karmazyn, B.; Gonin, R.; Hanna, M.P.

    1997-01-01

    Background. CT of the chest for suspected pulmonary metastases in adults is generally performed using a breath-hold technique. The results may not be applicable to young children in whom breath-holding may be impossible. Objective. Determine the effect of breathing on the accuracy of pulmonary metastasis detection by spiral CT (SCT). Materials and methods. Prior to euthanasia four anesthetized dogs with metastatic osteosarcoma underwent SCT with a collimation of 5 mm and a pitch of 2, during both induced breath-hold and normal quiet breathing. Images were reconstructed as contiguous 5-mm slices. Macroscopically evident metastases were noted at postmortem. Hard-copy SCT images were reviewed by ten radiologists, each of whom circled all suspected metastases. SCT images were compared with postmortem results to determine true and false positives. Results. The pathologist identified 132 macroscopically evident pulmonary metastases. For metastasis detection, there was no significant difference between breath-hold SCT and breathing SCT. Conclusion. In our animal model, SCT can be performed during normal resting breathing without significant loss of accuracy in the detection of pulmonary metastases. (orig.). With 3 tabs

  18. Low-dose CT imaging of a total hip arthroplasty phantom using model-based iterative reconstruction and orthopedic metal artifact reduction

    Energy Technology Data Exchange (ETDEWEB)

    Wellenberg, R.H.H.; Streekstra, G.J.; Maas, M. [Academic Medical Center, Department of Radiology, Amsterdam (Netherlands); Boomsma, M.F.; Osch, J.A.C. van [Department of Radiology, Zwolle (Netherlands); Vlassenbroek, A. [Philips Medical Systems, Brussels (Belgium); Milles, J. [Philips Medical Systems, Eindhoven (Netherlands); Edens, M.A. [Department of Innovation and Science, Zwolle (Netherlands); Slump, C.H. [University of Twente, MIRA Institute for Biomedical Technology and Technical Medicine, Enschede (Netherlands)

    2017-05-15

    To compare quantitative measures of image quality, in terms of CT number accuracy, noise, signal-to-noise-ratios (SNRs), and contrast-to-noise ratios (CNRs), at different dose levels with filtered-back-projection (FBP), iterative reconstruction (IR), and model-based iterative reconstruction (MBIR) alone and in combination with orthopedic metal artifact reduction (O-MAR) in a total hip arthroplasty (THA) phantom. Scans were acquired from high- to low-dose (CTDI{sub vol}: 40.0, 32.0, 24.0, 16.0, 8.0, and 4.0 mGy) at 120- and 140- kVp. Images were reconstructed using FBP, IR (iDose{sup 4} level 2, 4, and 6) and MBIR (IMR, level 1, 2, and 3) with and without O-MAR. CT number accuracy in Hounsfield Units (HU), noise or standard deviation, SNRs, and CNRs were analyzed. The IMR technique showed lower noise levels (p < 0.01), higher SNRs (p < 0.001) and CNRs (p < 0.001) compared with FBP and iDose{sup 4} in all acquisitions from high- to low-dose with constant CT numbers. O-MAR reduced noise (p < 0.01) and improved SNRs (p < 0.01) and CNRs (p < 0.001) while improving CT number accuracy only at a low dose. At the low dose of 4.0 mGy, IMR level 1, 2, and 3 showed 83%, 89%, and 95% lower noise values, a factor 6.0, 9.2, and 17.9 higher SNRs, and 5.7, 8.8, and 18.2 higher CNRs compared with FBP respectively. Based on quantitative analysis of CT number accuracy, noise values, SNRs, and CNRs, we conclude that the combined use of IMR and O-MAR enables a reduction in radiation dose of 83% compared with FBP and iDose{sup 4} in the CT imaging of a THA phantom. (orig.)

  19. Determination of the integrated CT number of the whole liver in patients with severe hepatitis

    International Nuclear Information System (INIS)

    Kumahara, Tadashi; Muto, Yasutoshi; Moriwaki, Hisataka; Yoshida, Takashi; Tomita, Eiichi

    1989-01-01

    A study was conducted to estimate the functional reserve of the liver of patients with severe hepatitis by computed tomography (CT), in particular employing the integrated CT number of the whole liver (ICTN). ICTN was calculated by integrating the product of 'area' times 'mean CT number' of the liver in each CT slice for the entire height of the liver. The following results were obtained: 1) In patients with fulminant hepatitis (FH) as well as those with subacute hepatitis (SAH), ICTN was found to be significantly lower as compared to that of patients with acute hepatitis (AH) or non-hepatic diseases. In addition, in FH and SAH patients, ICTN showed a larger degree of decrease when compared with such conventional parameters as either estimated liver volume or mean hepatic CT number. Thus, ICTN seems to more sensitively reflect the changes in functional reserve of the liver. 2) ICTN showed significant positive correlations with prothrombin time and plasma BCAA/AAA ratio, and a significant negative correlation with plasma methionine level. 3) Time course of changes in ICTN correlated well with the clinical features of severe hepatitis. In particular, patients with initial ICTN values above 20 l·HU/m 2 of body surface area showed significantly higher survival rate than those with initial ICTN below 20. In conclusion, ICTN well indicates the functional reserve of the liver, and is further suggested to be valuable as a parameter to predict the prognosis of patients with severe hepatitis. (author)

  20. A MRI-CT prostate registration using sparse representation technique

    Science.gov (United States)

    Yang, Xiaofeng; Jani, Ashesh B.; Rossi, Peter J.; Mao, Hui; Curran, Walter J.; Liu, Tian

    2016-03-01

    technique was validated with a clinical study of 10 patients. The accuracy of our approach was assessed using some identified landmarks in both MRI and CT images. Our proposed registration was compared with the current free-form-deformation (FFD)-based registration method. The accuracy of the proposed method was significantly higher than the commonly used FFD-based registration utilizing normalized mutual information (NMI). Conclusions: We have developed a new prostate MR-CT registration approach based on patch-deformation dictionary, demonstrated its clinical feasibility, and validated its accuracy with some identified landmarks. The proposed registration method may provide an accurate and robust means of estimating prostate-gland deformation between MRI and CT scans, and is therefore well-suited for a number of MR-targeted CT-based prostate radiotherapy.

  1. Accuracy of the detection of infratentorial stroke lesions using perfusion CT: an experimenter-blinded study

    International Nuclear Information System (INIS)

    Lee, In Hyouk; Kim, Myung Soon; Kim, Young Ju; Lee, Myeong Sub; You, Joshua H.; Lee, Ji Yong; Whang, Kum

    2010-01-01

    Although perfusion CT (PCT) for the detection of supratentorial stroke is well established, there is a dearth of evidence of its effectiveness in the detection of infratentorial stroke. Hence, this study compared sensitivity, specificity, and accuracy of PCT maps between infratentorial and supratentorial stroke lesions. One hundred patients with acute stroke who had successfully undergone near whole-brain PCT with the toggling table technique and follow-up MRI were included. Wilcoxon Mann-Whitney test was performed at P 0.01 in all PCT maps) between supratentorial and infratentorial stroke. Also, there was no remarkable difference in both sensitivity and specificity of PCT maps. This was the first study to investigate the accuracy of PCT with the toggling table technique in detection of infratentorial stroke lesions. Clinically, PCT is highly reliable and accurate in detecting infratentorial stroke lesions. (orig.)

  2. Accuracy of CT-guided percutaneous core needle biopsy for assessment of pediatric musculoskeletal lesions

    International Nuclear Information System (INIS)

    Hryhorczuk, Anastasia L.; Strouse, Peter J.; Biermann, J.S.

    2011-01-01

    CT-guided percutaneous core needle biopsy has been shown in adults to be an effective diagnostic tool for a large number of musculoskeletal malignancies. To characterize our experience with CT-guided percutaneous core needle biopsy of pediatric bone lesions and determine its utility in diagnosing pediatric osseous lesions, in a population where such lesions are commonly benign. From 2000 to 2009, 61 children underwent 63 CT-guided percutaneous biopsies. Radiological, pathological and clinical records were reviewed. Fourteen biopsies (22%) were performed on malignant lesions, while 49 biopsies (78%) were performed on benign lesions. Forty-nine of the 63 biopsies (78%) were adequate; these children underwent no further tissue sampling. Fourteen of the 63 biopsies (22%) were inadequate or non-conclusive. Of these patients, 12 underwent open biopsy. Retrospective analysis of percutaneous biopsies in these patients demonstrates that 9/12 provided clinically relevant information, and 4/12 patients received final diagnoses that confirmed initial core biopsy findings. No malignancies were diagnosed as benign on percutaneous biopsy. Overall, percutaneous core needle biopsy provided accurate diagnostic information in 84% (53/63) of biopsies. Our results demonstrate that CT-guided percutaneous biopsy is safe and beneficial in children. This study supports the use of CT-guided percutaneous core needle biopsy for primary diagnosis of pediatric bone lesions. (orig.)

  3. Coronary CT angiography using prospective ECG triggering. High diagnostic accuracy with low radiation dose

    International Nuclear Information System (INIS)

    Arnoldi, E.; Ramos-Duran, L.; Abro, J.A.; Costello, P.; Zwerner, P.L.; Schoepf, U.J.; Nikolaou, K.; Reiser, M.F.

    2010-01-01

    The purpose of this study was to evaluate the diagnostic performance of coronary CT angiography (coronary CTA) using prospective ECG triggering (PT) for the detection of significant coronary artery stenosis compared to invasive coronary angiography (ICA). A total of 20 patients underwent coronary CTA with PT using a 128-slice CT scanner (Definition trademark AS+, Siemens) and ICA. All coronary CTA studies were evaluated for significant coronary artery stenoses (≥50% luminal narrowing) by 2 observers in consensus using the AHA-15-segment model. Findings in CTA were compared to those in ICA. Coronary CTA using PT had 88% sensitivity in comparison to 100% with ICA, 95% to 88% specificity, 80% to 92% positive predictive value and 97% to 100% negative predictive value for diagnosing significant coronary artery stenosis on per segment per patient analysis, respectively. Mean effective radiation dose-equivalent of CTA was 2.6±1 mSv. Coronary CTA using PT enables non-invasive diagnosis of significant coronary artery stenosis with high diagnostic accuracy in comparison to ICA and is associated with comparably low radiation exposure. (orig.) [de

  4. Effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies

    International Nuclear Information System (INIS)

    Murase, Kenya; Nanjo, Takafumi; Ii, Satoshi; Miyazaki, Shohei; Hirata, Masaaki; Sugawara, Yoshifumi; Kudo, Masayuki; Sasaki, Kousuke; Mochizuki, Teruhito

    2005-01-01

    The purpose of this study was to investigate the effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies using multi-detector row CT (MDCT). Following the standard CT perfusion study protocol, continuous (cine) scans (1 s/rotation x 60 s) consisting of four 5 mm thick contiguous slices were performed using an MDCT scanner with a tube voltage of 80 kVp and a tube current of 200 mA. We generated the simulated images with tube currents of 50 mA, 100 mA and 150 mA by adding the corresponding noise to the raw scan data of the original image acquired above using a noise simulation tool. From the original and simulated images, we generated the functional images of cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transit time (MTT) in seven patients with cerebrovascular disease, and compared the correlation coefficients (CCs) between the perfusion parameter values obtained from the original and simulated images. The coefficients of variation (CVs) in the white matter were also compared. The CC values deteriorated with decreasing tube current. There was a significant difference between 50 mA and 100 mA for all perfusion parameters. The CV values increased with decreasing tube current. There were significant differences between 50 mA and 100 mA and between 100 mA and 150 mA for CBF. For CBV and MTT, there was also a significant difference between 150 mA and 200 mA. This study will be useful for understanding the effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies using MDCT, and for selecting the tube current

  5. Preoperative 4D CT Localization of Nonlocalizing Parathyroid Adenomas by Ultrasound and SPECT-CT.

    Science.gov (United States)

    Hinson, Andrew M; Lee, David R; Hobbs, Bradley A; Fitzgerald, Ryan T; Bodenner, Donald L; Stack, Brendan C

    2015-11-01

    To evaluate 4-dimensional (4D) computed tomography (CT) for the localization of parathyroid adenomas previously considered nonlocalizing on ultrasound and single-photon emission CT with CT scanning (SPECT-CT). To measure radiation exposure associated with 4D-CT and compared it with SPECT-CT. Case series with chart review. University tertiary hospital. Nineteen adults with primary hyperparathyroidism who underwent preoperative 4D CT from November 2013 through July 2014 after nonlocalizing preoperative ultrasound and technetium-99m SPECT-CT scans. Sensitivity, specificity, predictive values, and accuracy of 4D CT were evaluated. Nineteen patients (16 women and 3 men) were included with a mean age of 66 years (range, 39-80 years). Mean preoperative parathyroid hormone level was 108.5 pg/mL (range, 59.3-220.9 pg/mL), and mean weight of the excised gland was 350 mg (range, 83-797 mg). 4D CT sensitivity and specificity for localization to the patient's correct side of the neck were 84.2% and 81.8%, respectively; accuracy was 82.9%. The sensitivity for localizing adenomas to the correct quadrant was 76.5% and 91.5%, respectively; accuracy was 88.2%. 4D CT radiation exposure was significantly less than the radiation associated with SPECT-CT (13.8 vs 18.4 mSv, P = 0.04). 4D CT localizes parathyroid adenomas with relatively high sensitivity and specificity and allows for the localization of some adenomas not observed on other sestamibi-based scans. 4D CT was also associated with less radiation exposure when compared with SPECT-CT based on our study protocol. 4D CT may be considered as first- or second-line imaging for localizing parathyroid adenomas in the setting of primary hyperparathyroidism. © American Academy of Otolaryngology—Head and Neck Surgery Foundation 2015.

  6. Detection of pulmonary fat embolism with dual-energy CT: an experimental study in rabbits

    Energy Technology Data Exchange (ETDEWEB)

    Tang, Chun Xiang; Zhou, Chang Sheng; Zhao, Yan E.; Han, Zong Hong; Qi, Li; Zhang, Long Jiang; Lu, Guang Ming [Medical School of Nanjing University, Department of Medical Imaging, Jinling Hospital, Nanjing, Jiangsu (China); Schoepf, U.J. [Medical School of Nanjing University, Department of Medical Imaging, Jinling Hospital, Nanjing, Jiangsu (China); Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States); Mangold, Stefanie; Ball, B.D. [Medical University of South Carolina, Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Charleston, SC (United States)

    2017-04-15

    To evaluate the use of dual-energy CT imaging of the lung perfused blood volume (PBV) for the detection of pulmonary fat embolism (PFE). Dual-energy CT was performed in 24 rabbits before and 1 hour, 1 day, 4 days and 7 days after artificial induction of PFE via the right ear vein. CT pulmonary angiography (CTPA) and lung PBV images were evaluated by two radiologists, who recorded the presence, number, and location of PFE on a per-lobe basis. Sensitivity, specificity, and accuracy of CTPA and lung PBV for detecting PFE were calculated using histopathological evaluation as the reference standard. A total of 144 lung lobes in 24 rabbits were evaluated and 70 fat emboli were detected on histopathological analysis. The overall sensitivity, specificity and accuracy were 25.4 %, 98.6 %, and 62.5 % for CTPA, and 82.6 %, 76.0 %, and 79.2 % for lung PBV. Higher sensitivity (p < 0.001) and accuracy (p < 0.01), but lower specificity (p < 0.001), were found for lung PBV compared with CTPA. Dual-energy CT can detect PFE earlier than CTPA (all p < 0.01). Dual-energy CT provided higher sensitivity and accuracy in the detection of PFE as well as earlier detection compared with conventional CTPA in this animal model study. (orig.)

  7. Diagnostic accuracy of 18F-FDG-PET and PET/CT in patients with Ewing sarcoma family tumours: a systematic review and a meta-analysis

    International Nuclear Information System (INIS)

    Treglia, Giorgio; Salsano, Marco; Stefanelli, Antonella; Mattoli, Maria Vittoria; Giordano, Alessandro; Bonomo, Lorenzo

    2012-01-01

    To systematically review and meta-analyse literature data on the diagnostic performance of fluorine-18-fluorodeoxyglucose (FDG) positron emission tomography (PET) and positron emission tomography/computed tomography (PET/CT) in patients with Ewing sarcoma family tumours (ESFT). PubMed/MEDLINE, Embase and Scopus databases were searched for articles that evaluated FDG-PET and PET/CT in patients with ESFT from inception to 31 May 2011. Studies that fulfilled the three following criteria were included in the systematic review: FDG-PET or PET/CT performed in patients with ESFT; articles about the diagnostic accuracy of FDG-PET and PET/CT; sample size of at least 10 patients with ESFT were included. Studies in which there were sufficient data to reassess sensitivity and specificity of FDG-PET or PET/CT in ESFT were included in the meta-analysis, excluding duplicate publications. Finally, pooled sensitivity, pooled specificity and area under the receiver operating characteristic (ROC) curve of FDG-PET or PET/CT in ESFT were calculated. We found 13 studies comprising a total of 342 patients with ESFT. The main findings of the studies included are presented. The meta-analysis of five selected studies provided these results about FDG-PET and PET/CT in ESFT: pooled sensitivity: 96% (95% confidence interval [CI] 91-99%); pooled specificity: 92% (95% CI 87-96%); area under the ROC curve: 0.97. With regard to the staging and restaging of patients with ESFT, the sensitivity, specificity and accuracy of FDG-PET and PET/CT are high; the combination of FDG-PET or PET/CT with conventional imaging is a valuable tool for the staging and restaging of ESFT and has a relevant impact on the treatment strategy plan. (orig.)

  8. Assessing the Dosimetric Accuracy of Magnetic Resonance-Generated Synthetic CT Images for Focal Brain VMAT Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Paradis, Eric, E-mail: eparadis@umich.edu [Department of Radiation Oncology, University of Michigan Hospital and Health Systems, Ann Arbor, Michigan (United States); Cao, Yue [Department of Radiation Oncology, University of Michigan Hospital and Health Systems, Ann Arbor, Michigan (United States); Department of Radiology, University of Michigan Hospital and Health Systems, Ann Arbor, Michigan (United States); Department of Biomedical Engineering, University of Michigan Hospital and Health Systems, Ann Arbor, Michigan (United States); Lawrence, Theodore S.; Tsien, Christina; Feng, Mary; Vineberg, Karen; Balter, James M. [Department of Radiation Oncology, University of Michigan Hospital and Health Systems, Ann Arbor, Michigan (United States)

    2015-12-01

    Purpose: The purpose of this study was to assess the dosimetric accuracy of synthetic CT (MRCT) volumes generated from magnetic resonance imaging (MRI) data for focal brain radiation therapy. Methods and Materials: A study was conducted in 12 patients with gliomas who underwent both MR and CT imaging as part of their simulation for external beam treatment planning. MRCT volumes were generated from MR images. Patients' clinical treatment planning directives were used to create 12 individual volumetric modulated arc therapy (VMAT) plans, which were then optimized 10 times on each of their respective CT and MRCT-derived electron density maps. Dose metrics derived from optimization criteria, as well as monitor units and gamma analyses, were evaluated to quantify differences between the imaging modalities. Results: Mean differences between planning target volume (PTV) doses on MRCT and CT plans across all patients were 0.0% (range: −0.1 to 0.2%) for D{sub 95%}; 0.0% (−0.7 to 0.6%) for D{sub 5%}; and −0.2% (−1.0 to 0.2%) for D{sub max}. MRCT plans showed no significant changes in monitor units (−0.4%) compared to CT plans. Organs at risk (OARs) had average D{sub max} differences of 0.0 Gy (−2.2 to 1.9 Gy) over 85 structures across all 12 patients, with no significant differences when calculated doses approached planning constraints. Conclusions: Focal brain VMAT plans optimized on MRCT images show excellent dosimetric agreement with standard CT-optimized plans. PTVs show equivalent coverage, and OARs do not show any overdose. These results indicate that MRI-derived synthetic CT volumes can be used to support treatment planning of most patients treated for intracranial lesions.

  9. Dose reduction for CT in children with cystic fibrosis: is it feasible to reduce the number of images per scan?

    International Nuclear Information System (INIS)

    Jong, Pim A. de; Tiddens, Harm A.W.M.; Nakano, Yasutaka; Lequin, Maarten H.

    2006-01-01

    Reducing the dose for each CT scan is important for children with cystic fibrosis (CF). To determine whether the number of CT images and therefore the dose per CT scan could be reduced without any significant loss of information in children with CF. A cohort of children with CF was followed with biennial surveillance CT scans, obtained in inspiration after a voluntary breath-hold as 1-mm thick images at 10-mm intervals from lung apex to base. A random set of 20 baseline CT scans and 10 follow-up CT scans were blinded. Sets of every image (10-mm intervals), every second image (20-mm intervals), every third image (30-mm intervals) and a selection of three and five images were scored randomly using a published CT scoring system by one experienced observer. The 20 subjects were 10 years of age with a range of 3.7-17.6 years at baseline. Fewer CT images resulted in a significantly lower (less abnormal) CT score and the number of patients positive for abnormalities decreased subsequently. At intervals greater than 20 mm no significant change in CT score over 2 years could be detected, while the CT scores at 10-mm (P=0.02) and 20-mm (P=0.02) intervals worsened significantly. A reduction in the number of inspiratory CT images by increasing the interval between images to greater than 10 mm is not a valid option for radiation dose reduction in children with CF. (orig.)

  10. Diagnostic accuracy of sub-mSv prospective ECG-triggering cardiac CT in young infant with complex congenital heart disease.

    Science.gov (United States)

    Gao, Wei; Zhong, Yu Min; Sun, Ai Min; Wang, Qian; Ouyang, Rong Zhen; Hu, Li Wei; Qiu, Han Sheng; Wang, Shi Yu; Li, Jian Ying

    2016-06-01

    To explore the clinical value and evaluate the diagnostic accuracy of sub-mSv low-dose prospective ECG-triggering cardiac CT (CCT) in young infants with complex congenital heart disease (CHD). A total of 102 consecutive infant patients (53 boys and 49 girls with mean age of 2.9 ± 2.4 m and weight less than 5 kg) with complex CHD were prospectively enrolled. Scans were performed on a 64-slice high definition CT scanner with low dose prospective ECG-triggering mode and reconstructed with 80 % adaptive statistical iterative reconstruction algorithm. All studies were performed during free breathing with sedation. The subjective image quality was evaluated by 5-point grading scale and interobserver variability was calculated. The objective image noise (standard deviation, SD) and contrast to noise ratio (CNR) was calculated. The effective radiation dose from the prospective ECG-triggering mode was recorded and compared with the virtual conventional retrospective ECG-gating mode. The detection rate for the origin of coronary artery was calculated. All patients also underwent echocardiography before CCT examination. 81 patients had surgery and their preoperative CCT and echocardiography findings were compared with the surgical results and sensitivity, specificity, positive and negative predictive values and accuracy were calculated for separate cardiovascular anomalies. Heart rates were 70-161 beats per minute (bpm) with mean value of 129.19 ± 14.52 bpm. The effective dose of 0.53 ± 0.15 mSv in the prospective ECG-triggering cardiac CT was lower than the calculated value in a conventional retrospective ECG-gating mode (2.00 ± 0.35 mSv) (p ECG-triggering CCT with sub-mSv effective dose provides excellent imaging quality and high diagnostic accuracy for young infants with complex CHD.

  11. Diagnostic accuracy of 18F-FDG-PET and PET/CT in the differential diagnosis between malignant and benign pleural lesions: a systematic review and meta-analysis.

    Science.gov (United States)

    Treglia, Giorgio; Sadeghi, Ramin; Annunziata, Salvatore; Lococo, Filippo; Cafarotti, Stefano; Bertagna, Francesco; Prior, John O; Ceriani, Luca; Giovanella, Luca

    2014-01-01

    To systematically review and meta-analyze published data about the diagnostic accuracy of fluorine-18-fluorodeoxyglucose ((18)F-FDG) positron emission tomography (PET) and PET/computed tomography (CT) in the differential diagnosis between malignant and benign pleural lesions. A comprehensive literature search of studies published through June 2013 regarding the diagnostic performance of (18)F-FDG-PET and PET/CT in the differential diagnosis of pleural lesions was carried out. All retrieved studies were reviewed and qualitatively analyzed. Pooled sensitivity, specificity, positive and negative likelihood ratio (LR+ and LR-) and diagnostic odds ratio (DOR) of (18)F-FDG-PET or PET/CT in the differential diagnosis of pleural lesions on a per-patient-based analysis were calculated. The area under the summary receiver operating characteristic curve (AUC) was calculated to measure the accuracy of these methods. Subanalyses considering device used (PET or PET/CT) were performed. Sixteen studies including 745 patients were included in the systematic review. The meta-analysis of 11 selected studies provided the following results: sensitivity 95% (95% confidence interval [95%CI]: 92-97%), specificity 82% (95%CI: 76-88%), LR+ 5.3 (95%CI: 2.4-11.8), LR- 0.09 (95%CI: 0.05-0.14), DOR 74 (95%CI: 34-161). The AUC was 0.95. No significant improvement of the diagnostic accuracy considering PET/CT studies only was found. (18)F-FDG-PET and PET/CT demonstrated to be accurate diagnostic imaging methods in the differential diagnosis between malignant and benign pleural lesions; nevertheless, possible sources of false-negative and false-positive results should be kept in mind. Copyright © 2014 AUR. Published by Elsevier Inc. All rights reserved.

  12. Results and complications of CT-guided biopsy with CT fluoroscopy

    International Nuclear Information System (INIS)

    Saika, Yoshinori; Ogura, Yasuharu; Doi, Kenji; Misaki, Toshimasa; Shimizu, Masashi; Narabayashi, Isamu

    2002-01-01

    We studied the results and complications of CT-guided biopsy with CT fluoroscopy performed 66 lesions in 64 patients from March 1999 to February 2001. In addition to the conventional procedure of CT-guided biopsy, we use CT fluoroscopy for confirmation of the location of the tip of the biopsy needle and the accurate contact, in some cases, at the time of puncturing. Examination results showed malignancy in 36 lesions and benign findings in 30 lesions. The sensitivity was 85.7%, specificity was 100.0%, and accuracy was 90.9%. Pneumothorax occurred in 20 out of 64 patients (31.3%). In a study on 26 small lesions (≤2 cm) in 25 patients, the sensitivity was 81.8%, specificity was 100.0%, accuracy was 92.3%. Pneumothorax occurred in 12 out of 25 patients (48.0%), more frequently than in patients with large lesions. In a study on 6 false negative cases, they tended to be intrapulmonary on location, small in diameter, and far from the skin puncture point. Examination results were satisfactory, especially in terms of accuracy in small lesions (≤2 cm). However, when the lesions were small, pneumothorax occurred frequently (48.0%). (author)

  13. Detection of pulmonary metastases with pathological correlation: effect of breathing on the accuracy of spiral CT. Editor`s note

    Energy Technology Data Exchange (ETDEWEB)

    Coakley, F.V. [Abdominal Imaging Section, Department of Radiology, University of California, San Francisco, CA (United States); Cohen, M.D. [Department of Radiology, Riley Hospital for Children, Indianapolis, IN (United States); Waters, D.J. [Department of Veterinary Clinical Sciences, Purdue University, West Lafayette, IN (United States); Davis, M.M. [Department of Pathology, Riley Hospital for Children, Indianapolis, IN (United States); Karmazyn, B. [Department of Radiology, Riley Hospital for Children, Indianapolis, IN (United States); Gonin, R. [Division of Biostatistics, Department of Medicine, Riley Hospital for Children, Indianapolis, IN (United States); Hanna, M.P. [Division of Biostatistics, Department of Medicine, Riley Hospital for Children, Indianapolis, IN (United States)

    1997-07-01

    Background. CT of the chest for suspected pulmonary metastases in adults is generally performed using a breath-hold technique. The results may not be applicable to young children in whom breath-holding may be impossible. Objective. Determine the effect of breathing on the accuracy of pulmonary metastasis detection by spiral CT (SCT). Materials and methods. Prior to euthanasia four anesthetized dogs with metastatic osteosarcoma underwent SCT with a collimation of 5 mm and a pitch of 2, during both induced breath-hold and normal quiet breathing. Images were reconstructed as contiguous 5-mm slices. Macroscopically evident metastases were noted at postmortem. Hard-copy SCT images were reviewed by ten radiologists, each of whom circled all suspected metastases. SCT images were compared with postmortem results to determine true and false positives. Results. The pathologist identified 132 macroscopically evident pulmonary metastases. For metastasis detection, there was no significant difference between breath-hold SCT and breathing SCT. Conclusion. In our animal model, SCT can be performed during normal resting breathing without significant loss of accuracy in the detection of pulmonary metastases. (orig.). With 3 tabs.

  14. Perforated appendicitis: accuracy of CT diagnosis and correlation of CT findings with the length of hospital

    International Nuclear Information System (INIS)

    Siddiqui, H.A.; Afzal, S.

    2007-01-01

    To determine the sensitivity and specificity of CT findings in the differentiation of perforated from nonperforated appendicitis and correlate CT diagnosis with the length of hospital stay. The study included 70 patients who presented with right lower quadrant abdominal pain and underwent preoperative CT scan followed by appendectomy. Patients were divided into two groups of having perforated and nonperforated appendicitis on the basis of CT scan findings. The surgical and pathological reports combined were considered the reference standard for the diagnosis of perforated appendicitis. Various CT scan findings and average duration of hospital stay in days was compared by t-test. Twenty-six (37%) of 70 patients had perforated appendicitis. It was correctly identified on pre-operative CT scan in 18 patients. There were 18 true positive diagnoses, 43 true negative diagnoses, 1 false positive diagnosis and 8 false negative diagnoses which yielded a sensitivity of 69%, specificity of 97%, positive predictive value of 94% and negative predictive value of 84%. Mean length of hospital stay in perforated group was 6.3 days and 2.9 days in nonperforated group. Severe periappendiceal inflammation, periappendiceal and or abdominopelvic fluid and abscess were significantly associated with perforated appendicitis and with a significant longer hospital stay (p <.001). CT scan is 69% sensitive and 97% specific for the diagnosis of perforated appendicitis and constellation of CT findings can be used to select patients with perforated appendicitis for initial non-operative management. Presence of CT signs of significant appendiceal inflammation is independent predictor of longer hospital stay. (author)

  15. Low tube voltage and low contrast material volume cerebral CT angiography

    International Nuclear Information System (INIS)

    Luo, Song; Zhang, Long Jiang; Lu, Guang Ming; Meinel, Felix G.; McQuiston, Andrew D.; Zhou, Chang Sheng; Qi, Li; Schoepf, U.J.

    2014-01-01

    To evaluate the image quality, radiation dose and diagnostic accuracy of low kVp and low contrast material volume cerebral CT angiography (CTA) in intracranial aneurysm detection. One hundred twenty patients were randomly divided into three groups (n = 40 for each): Group A, 70 ml iodinated contrast agent/120 kVp; group B, 30 ml/100 kVp; group C, 30 ml/80 kVp. The CT numbers, noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured in the internal carotid artery (ICA) and middle cerebral artery (MCA). Subjective image quality was evaluated. For patients undergoing DSA, diagnostic accuracy of CTA was calculated with DSA as reference standard and compared. CT numbers of ICA and MCA were higher in groups B and C than in group A (P < 0.01). SNR and CNR in groups A and B were higher than in group C (both P < 0.05). There was no difference in subjective image quality among the three groups (P = 0.939). Diagnostic accuracy for aneurysm detection among these groups had no statistical difference (P = 1.00). Compared with group A, the radiation dose of groups B and C was decreased by 45 % and 74 %. Cerebral CTA at 100 or 80 kVp using 30 ml contrast agent can obtain diagnostic image quality with a low radiation dose while maintaining the same diagnostic accuracy for aneurysm detection. (orig.)

  16. Accuracy limits for the determination of cortical width and density: the influence of object size and CT imaging parameters

    International Nuclear Information System (INIS)

    Prevrhal, S.; Engelke, K.; Kalender, W.A.

    1999-01-01

    In this study we analysed the accuracy of computed tomography (CT) measurements in assessing cortical bone. We determined the dependency of thickness and density measurements on the true width and density of the cortex and on the spatial resolution in the CT images using two optimized segmentation methods. As a secondary goal, we assessed the ability of CT to reflect small changes in cortical thickness. Two different bone-mimicking phantoms with varying cortical thickness were scanned with single-slice CT on a Somatom Plus 4 scanner. Images were reconstructed with both a standard and a high-resolution convolution kernel. Two special operator-independent segmentation methods were used to automatically detect the edges of the cortical shell. We measured cortical thickness and density and compared the phantom measurements with theoretical computations by simulating a cross-sectional shape of the cortical shell. Based on the simulations, we calculated CT's power to detect small changes in cortical thickness. Simulations and phantom measurements were in very good agreement. Cortical thickness could be measured with an error of less than 10% if the true thickness was larger than 0.9 (0.7) mm for the standard (high-resolution) kernel which is close to the full width at half maximum (FWHM) of the point spread functions for these kernels and our scanner. Density measurements yielded errors of less than 10% for true cortical thickness values above two to three times the FWHM corresponding to 2.5 (2) mm in our case. The simulations showed that a 10% change in cortical width would not be detected with satisfying probability in bones with a cortical shell thinner than 1.2 mm. An accurate determination of the cortical thickness is limited to bones with a thickness higher than the FWHM of the scanner's point spread function. Therefore, the use of a high-resolution reconstruction kernel is crucial. Cortical bone mineral density can only be measured accurately in bones two to three

  17. Accuracy of the detection of infratentorial stroke lesions using perfusion CT: an experimenter-blinded study

    Energy Technology Data Exchange (ETDEWEB)

    Lee, In Hyouk; Kim, Myung Soon; Kim, Young Ju; Lee, Myeong Sub [Yonsei University, Department of Radiology, Wonju College of Medicine, Wonju (Korea, Republic of); You, Joshua H. [Center for Health, Wellness, Fitness, Prevention, and Healing Across the Life Span, Department of Physical Therapy, Graduate School of Rehabilitation Science, Yonsei University, Center for Movement Impairment Solutions, Wonju City (Korea, Republic of); Lee, Ji Yong [Yonsei University, Department of Neurology, Wonju College of Medicine, Wonju (Korea, Republic of); Whang, Kum [Yonsei University, Department of Neurosurgery, Wonju College of Medicine, Wonju (Korea, Republic of)

    2010-12-15

    Although perfusion CT (PCT) for the detection of supratentorial stroke is well established, there is a dearth of evidence of its effectiveness in the detection of infratentorial stroke. Hence, this study compared sensitivity, specificity, and accuracy of PCT maps between infratentorial and supratentorial stroke lesions. One hundred patients with acute stroke who had successfully undergone near whole-brain PCT with the toggling table technique and follow-up MRI were included. Wilcoxon Mann-Whitney test was performed at P < 0.01. There was no significant statistical difference in the accuracy (91.79% vs. 93.23% in regional cerebral blood volume; 92.26% vs. 95.31% in regional cerebral blood flow; 89.17% vs. 92.71% in mean transit time; 89.76% vs. 92.19% in time to peak; P > 0.01 in all PCT maps) between supratentorial and infratentorial stroke. Also, there was no remarkable difference in both sensitivity and specificity of PCT maps. This was the first study to investigate the accuracy of PCT with the toggling table technique in detection of infratentorial stroke lesions. Clinically, PCT is highly reliable and accurate in detecting infratentorial stroke lesions. (orig.)

  18. Congenital bronchopulmonary malformation: CT histopathological correlation.

    Science.gov (United States)

    Kyncl, Martin; Koci, Martin; Ptackova, Lea; Hornofova, Ludmila; Ondrej, Fabian; Snajdauf, Jiri; Pychova, Marcela

    2016-12-01

    This study evaluated the accuracy of postnatal computed tomography (CT) imaging in the identification of congenital bronchopulmonary malformation (BPM) in comparison with histopathological analysis. CT scans of prenatally diagnosed BPMs from 24 patients with available histology were analysed retrospectively. The CT images were reviewed blinded to histological findings by two radiologists. Specific diagnosis was assigned based on predetermined criteria. The accuracy of CT was evaluated. The agreement rate in CT diagnosis between two radiologists was 100%. In 75% the lesions were located in the lower lobes. An overlap of 71% in CT and histopathological diagnoses was reached. The least matching diagnosis was type 2 CPAM. Contrast enhanced chest CT is very accurate in characterizing the BPM spectrum and provides important information on lesion type and structure.

  19. SU-E-I-13: Evaluation of Metal Artifact Reduction (MAR) Software On Computed Tomography (CT) Images

    International Nuclear Information System (INIS)

    Huang, V; Kohli, K

    2015-01-01

    Purpose: A new commercially available metal artifact reduction (MAR) software in computed tomography (CT) imaging was evaluated with phantoms in the presence of metals. The goal was to assess the ability of the software to restore the CT number in the vicinity of the metals without impacting the image quality. Methods: A Catphan 504 was scanned with a GE Optima RT 580 CT scanner (GE Healthcare, Milwaukee, WI) and the images were reconstructed with and without the MAR software. Both datasets were analyzed with Image Owl QA software (Image Owl Inc, Greenwich, NY). CT number sensitometry, MTF, low contrast, uniformity, noise and spatial accuracy were compared for scans with and without MAR software. In addition, an in-house made phantom was scanned with and without a stainless steel insert at three different locations. The accuracy of the CT number and metal insert dimension were investigated as well. Results: Comparisons between scans with and without MAR algorithm on the Catphan phantom demonstrate similar results for image quality. However, noise was slightly higher for the MAR algorithm. Evaluation of the CT number at various locations of the in-house made phantom was also performed. The baseline HU, obtained from the scan without metal insert, was compared to scans with the stainless steel insert at 3 different locations. The HU difference between the baseline scan versus metal scan was improved when the MAR algorithm was applied. In addition, the physical diameter of the stainless steel rod was over-estimated by the MAR algorithm by 0.9 mm. Conclusion: This work indicates with the presence of metal in CT scans, the MAR algorithm is capable of providing a more accurate CT number without compromising the overall image quality. Future work will include the dosimetric impact on the MAR algorithm

  20. Dual-energy CT angiography in peripheral arterial occlusive disease - accuracy of maximum intensity projections in clinical routine and subgroup analysis

    International Nuclear Information System (INIS)

    Kau, Thomas; Eicher, Wolfgang; Reiterer, Christian; Niedermayer, Martin; Rabitsch, Egon; Hausegger, Klaus A.; Senft, Birgit

    2011-01-01

    To evaluate the accuracy of dual-energy CT angiography (DE-CTA) maximum intensity projections (MIPs) in symptomatic peripheral arterial occlusive disease (PAOD). In 58 patients, DE-CTA of the lower extremities was performed on dual-source CT. In a maximum of 35 arterial segments, severity of the most stenotic lesion was graded (<10%, 10-49% and 50-99% luminal narrowing or occlusion) independently by two radiologists, with DSA serving as the reference standard. In DSA, 52.3% of segments were significantly stenosed or occluded. Agreement of DE-CTA MIPs with DSA was good in the aorto-iliac and femoro-popliteal regions (κ = 0.72; κ = 0.66), moderate in the crural region (κ = 0.55), slight in pedal arteries (κ = 0.10) and very good in bypass segments (κ = 0.81). Accuracy was 88%, 78%, 74%, 55% and 82% for the respective territories and moderate (75%) overall, with good sensitivity (84%) and moderate specificity (67%). Sensitivity and specificity was 82% and 76% in claudicants and 84% and 61% in patients with critical limb ischaemia. While correlating well with DSA above the knee, accuracy of DE-CTA MIPs appeared to be moderate in the calf and largely insufficient in calcified pedal arteries, especially in patients with critical limb ischaemia. (orig.)

  1. Diagnostic accuracy of non-contrast abdominal CT scans performed as follow-up for patients with an established cancer diagnosis: a retrospective study.

    Science.gov (United States)

    Semaan, Hassan; Bazerbashi, Mohamad F; Siesel, Geoffrey; Aldinger, Paul; Obri, Tawfik

    2018-03-01

    To determine the accuracy and non-detection rate of cancer related findings (CRFs) on follow-up non-contrast-enhanced CT (NECT) versus contrast-enhanced CT (CECT) images of the abdomen in patients with a known cancer diagnosis. A retrospective review of 352 consecutive CTs of the abdomen performed with and without IV contrast between March 2010 and October 2014 for follow-up of cancer was included. Two radiologists independently assessed the NECT portions of the studies. The reader was provided the primary cancer diagnosis and access to the most recent prior NECT study. The accuracy and non-detection rates were determined by comparing our results to the archived reports as a gold standard. A total of 383 CRFs were found in the archived reports of the 352 abdominal CTs. The average non-detection rate for the NECTs compared to the CECTs was 3.0% (11.5/383) with an accuracy of 97.0% (371.5/383) in identifying CRFs. The most common findings missed were vascular thrombosis with a non-detection rate of 100%. The accuracy for non-vascular CRFs was 99.1%. Follow-up NECT abdomen studies are highly accurate in the detection of CRFs in patients with an established cancer diagnosis, except in cases where vascular involvement is suspected.

  2. CT-based virtual tracheobronchoscopy in children - comparison with axial CT and multiplanar reconstruction: preliminary results

    International Nuclear Information System (INIS)

    Sorantin, Erich; Lindbichler, Franz; Eber, Ernst; Schimpl, Guenther

    2002-01-01

    Background: 3D post-processing of spiral-CT (S-CT) data using perspective projection allows the generation of virtual views similar to endoscopy. Objective: To evaluate whether simultaneous reading of axial S-CT, multiplanar reconstruction (MPR) and virtual tracheobronchoscopy (VTB) is more precise and accurate than reading of axial S-CT and MPR alone in paediatric patients. Materials and methods: S-CT studies of 15 symptomatic and 4 normal patients were investigated. Two radiologists independently read two sets of images for airway abnormalities: first axial CT and MPR, followed by axial CT, MPR and VTB. A final decision was later made by consensus. All results were compared to fibre-optic bronchoscopy (FTB). Interobserver agreement was used as an indicator of precision for the display technique used. Results: At reading of axial S-CT and MPR an interobserver agreement of 89.5% (κ=0.776, P<0.00103) was found. Based on the consensus decision, a diagnostic accuracy of 89.5% at a sensitivity 86.6% and specificity of 100% (κ=0.776, 95% CI 0.491-1.062, P<0.00103) was achieved. At reporting on axial S-CT, MPR and VTB, all cases were classified correctly by both readers, indicating 100% accuracy, interobserver agreement, sensitivity and specificity (κ=1.00, 95% CI 1.0-1.0, P<0.000258). Conclusions: The simultaneous display of axial S-CT, MPR and VTB raises the precision, accuracy and sensitivity of radiological reports. (orig.)

  3. MRI-Based Computed Tomography Metal Artifact Correction Method for Improving Proton Range Calculation Accuracy

    International Nuclear Information System (INIS)

    Park, Peter C.; Schreibmann, Eduard; Roper, Justin; Elder, Eric; Crocker, Ian; Fox, Tim; Zhu, X. Ronald; Dong, Lei; Dhabaan, Anees

    2015-01-01

    Purpose: Computed tomography (CT) artifacts can severely degrade dose calculation accuracy in proton therapy. Prompted by the recently increased popularity of magnetic resonance imaging (MRI) in the radiation therapy clinic, we developed an MRI-based CT artifact correction method for improving the accuracy of proton range calculations. Methods and Materials: The proposed method replaces corrupted CT data by mapping CT Hounsfield units (HU number) from a nearby artifact-free slice, using a coregistered MRI. MRI and CT volumetric images were registered with use of 3-dimensional (3D) deformable image registration (DIR). The registration was fine-tuned on a slice-by-slice basis by using 2D DIR. Based on the intensity of paired MRI pixel values and HU from an artifact-free slice, we performed a comprehensive analysis to predict the correct HU for the corrupted region. For a proof-of-concept validation, metal artifacts were simulated on a reference data set. Proton range was calculated using reference, artifactual, and corrected images to quantify the reduction in proton range error. The correction method was applied to 4 unique clinical cases. Results: The correction method resulted in substantial artifact reduction, both quantitatively and qualitatively. On respective simulated brain and head and neck CT images, the mean error was reduced from 495 and 370 HU to 108 and 92 HU after correction. Correspondingly, the absolute mean proton range errors of 2.4 cm and 1.7 cm were reduced to less than 2 mm in both cases. Conclusions: Our MRI-based CT artifact correction method can improve CT image quality and proton range calculation accuracy for patients with severe CT artifacts

  4. SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction

    International Nuclear Information System (INIS)

    Lin, T; Hamacher, K

    2016-01-01

    Purpose: This study utilizes a commercial PET/CT phantom to investigate two specific properties of a PET/CT system: the alignment accuracy of PET images with those from CT used for attenuation correction and the accuracy of this correction in PET images. Methods: A commercial PET/CT phantom consisting of three aluminum rods, two long central cylinders containing uniform activity, and attenuating materials such as air, water, bone and iodine contrast was scanned using a standard PET/CT protocol. Images reconstructed with 2 mm slice thickness and a 512 by 512 matrix were obtained. The center of each aluminum rod in the PET and CT images was compared to evaluate alignment accuracy. ROIs were drawn on transaxial images of the central rods at each section of attenuating material to determine the corrected activity (in BQML). BQML values were graphed as a function of slice number to provide a visual representation of the attenuation-correction throughout the whole phantom. Results: Alignment accuracy is high between the PET and CT images. The maximum deviation between the two in the axial plane is less than 1.5 mm, which is less than the width of a single pixel. BQML values measured along different sections of the large central rods are similar among the different attenuating materials except iodine contrast. Deviation of BQML values in the air and bone sections from the water section is less than 1%. Conclusion: Accurate alignment of PET and CT images is critical to ensure proper calculation and application of CT-based attenuation correction. This study presents a simple and quick method to evaluate the two with a single acquisition. As the phantom also includes spheres of increasing diameter, this could serve as a straightforward means to annually evaluate the status of a modern PET/CT system.

  5. SU-F-I-59: Quality Assurance Phantom for PET/CT Alignment and Attenuation Correction

    Energy Technology Data Exchange (ETDEWEB)

    Lin, T; Hamacher, K [Columbia University Medical Center, New York, NY (United States)

    2016-06-15

    Purpose: This study utilizes a commercial PET/CT phantom to investigate two specific properties of a PET/CT system: the alignment accuracy of PET images with those from CT used for attenuation correction and the accuracy of this correction in PET images. Methods: A commercial PET/CT phantom consisting of three aluminum rods, two long central cylinders containing uniform activity, and attenuating materials such as air, water, bone and iodine contrast was scanned using a standard PET/CT protocol. Images reconstructed with 2 mm slice thickness and a 512 by 512 matrix were obtained. The center of each aluminum rod in the PET and CT images was compared to evaluate alignment accuracy. ROIs were drawn on transaxial images of the central rods at each section of attenuating material to determine the corrected activity (in BQML). BQML values were graphed as a function of slice number to provide a visual representation of the attenuation-correction throughout the whole phantom. Results: Alignment accuracy is high between the PET and CT images. The maximum deviation between the two in the axial plane is less than 1.5 mm, which is less than the width of a single pixel. BQML values measured along different sections of the large central rods are similar among the different attenuating materials except iodine contrast. Deviation of BQML values in the air and bone sections from the water section is less than 1%. Conclusion: Accurate alignment of PET and CT images is critical to ensure proper calculation and application of CT-based attenuation correction. This study presents a simple and quick method to evaluate the two with a single acquisition. As the phantom also includes spheres of increasing diameter, this could serve as a straightforward means to annually evaluate the status of a modern PET/CT system.

  6. Evaluation of the accuracy of ventricular volume measurement by ultrafast CT

    International Nuclear Information System (INIS)

    Cui Wei; Dai Ruping; Guo Yuyin

    1997-01-01

    The authors evaluated the accuracy of ventricular volume measured by ultrafast CT (UFCT); and (2) compared the value of ventricular volume derived from long- and short-axis view. Fourteen human left ventricular casts and 15 right ventricular casts were scanned by Imatron C-150 scanner along both the long- and short-axis. The scan protocol was similar to that used in vivo. Eight 7 mm-thick slices were obtained from each cast for both long- and short-axis views. Ventricular volume was determined by the modified Simpson's rule provided by Inamtron Inc. The actual volumes of the ventricular casts were determined by the amount of water displacement by the cast. The actual volumes for left and right ventricles were 55.57 +- 28.91 ml and 64.23 +- 24.51 ml, respectively, the left and right ventricular volumes determined by UFCT were 66.50 +- 33.04 ml and 76.47 +-28.70 ml from long-axis view, and 60.36 +- 29.90 ml and 75.36 +- 28.73 ml from short-axis view, respectively. The measurements by UFCT were significantly greater than the actual volumes of the casts, both for the left and right ventricles (P 0.990). Both left and right ventricular volumes can be determined by UFCT with identical accuracy for both long- and short-axis views in calculating ventricular volume; however, overestimation of ventricular volume by UFCT should be noted

  7. Accuracy and role of contrast-enhanced CT in diagnosis and surgical planning in 88 soft tissue tumours of extremities

    International Nuclear Information System (INIS)

    Verga, Lucia; Robiati, Sara; De Marchi, Armanda; Martorano, Domenico; Faletti, Carlo; Brach del Prever, Elena Maria; Linari, Alessandra; Boffano, Michele; Piana, Raimondo

    2016-01-01

    Soft tissue tumours (STT) require accurate diagnosis in order to identify potential malignancies. Preoperative planning is fundamental to avoid inadequate treatments. The role of contrast-enhanced computed tomography (CT) for local staging remains incompletely assessed. Aims of the study were to evaluate CT accuracy in discriminating active from aggressive tumours compared to histology and evaluate the role of CT angiography (CTA) in surgical planning. This retrospective cohort series of 88 cases from 1200 patients (7 %) was locally studied by contrast-enhanced CT and CTA in a referral centre: 74 malignant tumours, 14 benign lesions. Contrast-enhancement patterns and relationship of the mass with major vessels and bone were compared with histology on surgically excised samples. Sensitivity, specificity, positive and negative predictive values (PPV, NPV) were evaluated in discriminating active from aggressive tumours. Sensitivity in differentiating aggressive tumours from active lesions was 89 %, specificity 84 %, PPV 90 %, NPV 82 %. The relationship between mass and major vessels/bone was fundamental for surgical strategy respectively in 40 % and in 58 % of malignant tumours. Contrast-enhanced CT and CTA are effective in differentiating aggressive masses from active lesions in soft tissue and in depicting the relationship between tumour and adjacent bones and major vessels. (orig.)

  8. Accuracy of whole-body FDG-PET and FDG-PET/CT in M staging of nasopharyngeal carcinoma: A systematic review and meta-analysis

    International Nuclear Information System (INIS)

    Chang, Ming-Che; Chen, Jin-Hua; Liang, Ji-An; Yang, Kuang-Tao; Cheng, Kai-Yuan; Kao, Chia-Hung

    2013-01-01

    Background: A meta-analysis was conducted to evaluate the accuracy of whole-body positron emission tomography (PET) or PET/CT in M staging of nasopharyngeal carcinoma (NPC). Methods: Through a search of relevant English language studies from October 1996 to September 2011, pooled estimated sensitivity, specificity, positive likelihood ratios, negative likelihood ratios, and summary receiver operating characteristic (SROC) curves of whole-body PET or PET/CT in M staging of NPC were calculated. Results: Three PET and 5 PET/CT studies were identified. The pooled sensitivity, specificity, positive likelihood ratio, and negative likelihood ratio of FDG-PET or PET/CT were 0.83 (95% confidence interval [CI], 0.77–0.88), 0.97 (95% CI, 0.95–0.98), 23.38 (95% CI, 16.22–33.69), and 0.19 (95% CI, 0.13–0.25), respectively. The area under curve was 0.9764 and Q* index estimate was 0.9307 for FDG-PET or PET/CT. Conclusion: Current evidence confirms the good diagnostic performance of the whole-body FDG-PET or PET/CT in M staging of NPC

  9. The iPad as a mobile device for CT display and interpretation: diagnostic accuracy for identification of pulmonary embolism.

    Science.gov (United States)

    Johnson, Pamela T; Zimmerman, Stefan L; Heath, David; Eng, John; Horton, Karen M; Scott, William W; Fishman, Elliot K

    2012-08-01

    Recent software developments enable interactive, real-time axial, 2D and 3D CT display on an iPad by cloud computing from a server for remote rendering. The purpose of this study was to compare radiologists' interpretative performance on the iPad to interpretation on the conventional picture archive and communication system (PACS). Fifty de-identified contrast-enhanced CT exams performed for suspected pulmonary embolism were compiled as an educational tool to prepare our residents for night call. Two junior radiology attendings blindly interpreted the cases twice, one reader used the PACS first, and the other interpreted on the iPad first. After an interval of at least 2 weeks, the cases were reinterpreted in different order using the other display technique. Sensitivity, specificity, and accuracy for identification of pulmonary embolism were compared for each interpretation method. Pulmonary embolism was present in 25 patients, ranging from main pulmonary artery to subsegmental thrombi. Both readers interpreted 98 % of cases correctly regardless of display platform. There was no significant difference in sensitivity (98 vs 100 %, p = 1.0), specificity (98 vs 96 %, p = 1.0), or accuracy (98 vs 98 %, p = 1.0) for interpretation with the iPad vs the PACS, respectively. CT interpretation on an iPad enabled accurate identification of pulmonary embolism, equivalent to display on the PACS. This mobile device has the potential to expand radiologists' availability for consultation and expedite emergency patient management.

  10. Diagnostic accuracy of {sup 18}F-FDG-PET and PET/CT in patients with Ewing sarcoma family tumours: a systematic review and a meta-analysis

    Energy Technology Data Exchange (ETDEWEB)

    Treglia, Giorgio [Institute of Nuclear Medicine, Catholic University of the Sacred Heart, Department of Bioimaging and Radiological Sciences, Rome (Italy); Institute of Nuclear Medicine, Positron Emission Tomography Centre, Catholic University of the Sacred Heart, Department of Bioimaging and Radiological Sciences, Rome (Italy); Salsano, Marco; Stefanelli, Antonella; Mattoli, Maria Vittoria; Giordano, Alessandro [Institute of Nuclear Medicine, Catholic University of the Sacred Heart, Department of Bioimaging and Radiological Sciences, Rome (Italy); Bonomo, Lorenzo [Institute of Radiology, Catholic University of the Sacred Heart, Department of Bioimaging and Radiological Sciences, Rome (Italy)

    2012-03-15

    To systematically review and meta-analyse literature data on the diagnostic performance of fluorine-18-fluorodeoxyglucose (FDG) positron emission tomography (PET) and positron emission tomography/computed tomography (PET/CT) in patients with Ewing sarcoma family tumours (ESFT). PubMed/MEDLINE, Embase and Scopus databases were searched for articles that evaluated FDG-PET and PET/CT in patients with ESFT from inception to 31 May 2011. Studies that fulfilled the three following criteria were included in the systematic review: FDG-PET or PET/CT performed in patients with ESFT; articles about the diagnostic accuracy of FDG-PET and PET/CT; sample size of at least 10 patients with ESFT were included. Studies in which there were sufficient data to reassess sensitivity and specificity of FDG-PET or PET/CT in ESFT were included in the meta-analysis, excluding duplicate publications. Finally, pooled sensitivity, pooled specificity and area under the receiver operating characteristic (ROC) curve of FDG-PET or PET/CT in ESFT were calculated. We found 13 studies comprising a total of 342 patients with ESFT. The main findings of the studies included are presented. The meta-analysis of five selected studies provided these results about FDG-PET and PET/CT in ESFT: pooled sensitivity: 96% (95% confidence interval [CI] 91-99%); pooled specificity: 92% (95% CI 87-96%); area under the ROC curve: 0.97. With regard to the staging and restaging of patients with ESFT, the sensitivity, specificity and accuracy of FDG-PET and PET/CT are high; the combination of FDG-PET or PET/CT with conventional imaging is a valuable tool for the staging and restaging of ESFT and has a relevant impact on the treatment strategy plan. (orig.)

  11. Octree indexing of DICOM images for voxel number reduction and improvement of Monte Carlo simulation computing efficiency

    International Nuclear Information System (INIS)

    Hubert-Tremblay, Vincent; Archambault, Louis; Tubic, Dragan; Roy, Rene; Beaulieu, Luc

    2006-01-01

    The purpose of the present study is to introduce a compression algorithm for the CT (computed tomography) data used in Monte Carlo simulations. Performing simulations on the CT data implies large computational costs as well as large memory requirements since the number of voxels in such data reaches typically into hundreds of millions voxels. CT data, however, contain homogeneous regions which could be regrouped to form larger voxels without affecting the simulation's accuracy. Based on this property we propose a compression algorithm based on octrees: in homogeneous regions the algorithm replaces groups of voxels with a smaller number of larger voxels. This reduces the number of voxels while keeping the critical high-density gradient area. Results obtained using the present algorithm on both phantom and clinical data show that compression rates up to 75% are possible without losing the dosimetric accuracy of the simulation

  12. Bowel preparation for CT colonography

    International Nuclear Information System (INIS)

    Neri, Emanuele; Lefere, Philippe; Gryspeerdt, Stefaan; Bemi, Pietro; Mantarro, Annalisa; Bartolozzi, Carlo

    2013-01-01

    Bowel preparation represents an essential part of CT colonography, as the accuracy of the exam is strongly related to the adequacy of colonic cleansing, and a poor bowel preparation may compromise the diagnostic quality even despite optimization of all other acquisition parameters. Residual stool and fluid in the large bowel may affect the interpretation of the exam and may increase the number of false positives and false negatives. In this regard, the majority of patients having undergone CT colonography state that bowel preparation is the most unpleasant part. Unfortunately, to date no definite consensus has been reached about the ideal bowel preparation technique, and there is great variability in preparation strategies across diagnostic centers. The purpose of this review article is to describe the development and evolution of bowel preparation techniques in order to choose the best approach for optimizing the diagnostic quality of CT colonography in each patient

  13. Bowel preparation for CT colonography

    Energy Technology Data Exchange (ETDEWEB)

    Neri, Emanuele, E-mail: emanuele.neri@med.unipi.it [Diagnostic and Interventional Radiology, University of Pisa (Italy); Lefere, Philippe; Gryspeerdt, Stefaan [Department of Radiology, Stedelijk Ziekenhuis, Roeselare (Belgium); Bemi, Pietro; Mantarro, Annalisa; Bartolozzi, Carlo [Diagnostic and Interventional Radiology, University of Pisa (Italy)

    2013-08-15

    Bowel preparation represents an essential part of CT colonography, as the accuracy of the exam is strongly related to the adequacy of colonic cleansing, and a poor bowel preparation may compromise the diagnostic quality even despite optimization of all other acquisition parameters. Residual stool and fluid in the large bowel may affect the interpretation of the exam and may increase the number of false positives and false negatives. In this regard, the majority of patients having undergone CT colonography state that bowel preparation is the most unpleasant part. Unfortunately, to date no definite consensus has been reached about the ideal bowel preparation technique, and there is great variability in preparation strategies across diagnostic centers. The purpose of this review article is to describe the development and evolution of bowel preparation techniques in order to choose the best approach for optimizing the diagnostic quality of CT colonography in each patient.

  14. Accuracy of renal volume assessment in children by three-dimensional sonography; Nierenvolumetrie im Kindesalter: Genauigkeit der dreidimensionalen Sonographie im Vergleich zur konventionellen Sonographie und CT/MRT

    Energy Technology Data Exchange (ETDEWEB)

    Fritz, G.A.; Riccabona, M.; Bohdal, G. [Klinik fuer Radiologie, Universitaetsklinikum Graz (Austria); Quehenberger, F. [Inst. fuer medizinische Informatik, Statistik und Dokumentation, Universitaetsklinikum Graz (Austria)

    2003-04-01

    Purpose: Prospective evaluation of the accuracy of three-dimensional ultrasound (3DUS) to assess the renal parenchymal volume. Materials and Methods: CT, MRI, 2DUS and 3DUS were used to measure the renal volume in 40 patients (range: neonate to 17 years; mean age: 8.95 years). The 3DUS was determined with a Voluson 730 (Kretztechnik, GE) or an external 3D-system (EchoTech, GE) using electromagnetic positioning sensors attached to conventional 2DUS-equipment. The 2DUS volume was calculated with the ellipsoid equation and the 3DUS volume computed with the system integrated software. For CT and MRI, planimetric analysis was used to determine the renal parenchymal volume, whereby the dilated collecting system of a hydronephrosis was subtracted to obtain the real renal parenchymal volume. The results of 2DUS and 3DUS were compared to the results of CT and MRI, and inter- and intraobserver variabilities were calculated. Results: In 74 of 77 kidneys, the 3DUS study was of diagnostic quality. The accuracy of the 3DUS volumes compared well to the CT and MRI volumes with a mean difference of -1.8 {+-} 4.6% versus a mean difference of -2.4 {+-} 15.4% for 2DUS. In normal kidneys, the accuracy was -2.6 {+-} 4.4% for 3DUS and -3.8 {+-} 14.7% for 2DUS. In hydronephrosis, the accuracy was +4.0 {+-} 5.9% and +9.6 {+-} 21.3%, respectively, indicating that 3DUS is more accurate than 2DUS, particularly in kidneys with a dilated collecting system. Inter- and intraobserver variabilities were {+-} 7.3% and {+-} 5.3%. Conclusion: For assessing the renal parenchymal volume in children, 3DUS is feasible and comparable to CT and MRI. (orig.) [German] Studienziel: Prospektive Evaluation der Genauigkeit des drei-dimensionalen Ultraschalls (3DUS) bei der Volumetrie der kindlichen Niere im Vergleich zur CT/MRT. Methode: Bei 40 Patienten (Alter: 0 - 17 Jahre) mit einer klinisch indizierten CT/MRT des Abdomens oder des Harntrakts wurde zusaetzlich ein 2D- und 3DUS inklusive Volumetrie des

  15. Assessment of the accuracy and reliability of the Topcon CT80 non-contact tonometer.

    Science.gov (United States)

    Ogbuehi, Kelechi C

    2006-09-01

    The reliability of non-contact tonometers has been reported extensively in the literature. This study was designed to assess reliability of the new Topcon CT80 non-contact tonometer in normotensive subjects, using the Goldmann tonometer as the standard. The accuracy of the Topcon CT 80 non-contact tonometer was assessed by comparing its IOP assessments with those of the Goldmann applanation tonometer, on 60 right eyes of young healthy subjects with normal intraocular pressures. Each subject's intraocular pressure was assessed with each technique on two separate occasions, one week apart. The reliability of each technique was determined by the assessment of its inter-session repeatability using the Bland-Altman method. The 95 per cent limits of agreement for the two methods were also determined. No statistically significant difference was found between the average intraocular pressures measured with the two techniques (p > 0.05). The inter-session repeatability indices for the two techniques did not differ significantly (p > 0.05). The mean difference in intraocular measurements between the two techniques was 0.2 +/- 1.5 mmHg (mean +/- SD) and the 95 per cent limits of agreement were -3.14 and +2.74 mmHg, with the non-contact tonometer returning higher readings than the Goldmann tonometer. In this sample of normotensive subjects, the Topcon CT80 non-contact tonometer proved to be accurate and as reliable as the Goldmann tonometer in the assessment of intraocular pressure. Thus, it can be used as an objective clinical method for the assessment of normal intraocular pressure.

  16. Accuracy of bone SPECT/CT for identifying hardware loosening in patients who underwent lumbar fusion with pedicle screws

    Energy Technology Data Exchange (ETDEWEB)

    Hudyana, Hendrah; Maes, Alex [AZ Groeninge, Department of Nuclear Medicine, Kortrijk (Belgium); University Hospital Leuven, Department of Morphology and Medical Imaging, Leuven (Belgium); Vandenberghe, Thierry; Fidlers, Luc [AZ Groeninge, Department of Neurosurgery, Kortrijk (Belgium); Sathekge, Mike [University of Pretoria, Department of Nuclear Medicine, Pretoria (South Africa); Nicolai, Daniel [AZ Groeninge, Department of Nuclear Medicine, Kortrijk (Belgium); Wiele, Christophe van de [AZ Groeninge, Department of Nuclear Medicine, Kortrijk (Belgium); University Ghent, Department of Radiology and Nuclear Medicine, Ghent (Belgium)

    2016-02-15

    The aim of this retrospective study was to evaluate the accuracy of bone SPECT (single photon emission computed tomography)/CT (computed tomography) in diagnosing loosening of fixation material in patients with recurrent or persistent back pain that underwent lumbar arthrodesis with pedicle screws using surgery and clinical follow-up as gold standard A total of 48 patients (median age 49 years, range 21-81 years; 17 men) who had undergone lumbar spinal arthrodesis were included in this retrospective analysis. SPECT/CT results were compared to the gold standard of surgical evaluation or clinical follow-up. Positive SPECT/CT results were considered true positives if findings were confirmed by surgery or if clinical and other examinations were completely consistent with the positive SPECT/CT finding. They were considered false positives if surgical evaluation did not find any loose pedicle screws or if symptoms subsided with non-surgical therapy. Negative SPECT/CT scans were considered true negatives if symptoms either improved without surgical intervention or remained stable over a minimum follow-up period of 6 months. Negative SPECT/CT scans were determined to be false negatives if surgery was still required and loosening of material was found. The median length of time from primary surgery to bone SPECT/CT referral was 29.5 months (range 12-192 months). Median follow-up was 18 months (range 6-57) for subjects who did not undergo surgery. Thirteen of the 48 patients were found to be positive for loosening on bone SPECT/CT. Surgical evaluation (8 patients) and clinical follow-up (5 patients) showed that bone SPECT/CT correctly predicted loosening in 9 of 13 patients, while it falsely diagnosed loosening in 4 patients. Of 35 negative bone SPECT/CT scans, 12 were surgically confirmed. In 18 patients, bone SPECT/CT revealed lesions that could provide an alternative explanation for the symptoms of pain (active facet degeneration in 14 patients, and disc and sacroiliac

  17. The Value of Calcium-scoring CT for Ischemic Cardiovascular Disease Screening

    International Nuclear Information System (INIS)

    Oh, Jung Hoan; An, Sung Min

    2009-01-01

    The cardiovascular disease has been known as a common cause of death for a long time in the west. The eating habits of Asia, including Korea, have changed recently, so that this disease is also a problem in Asia now. Annual Report on the Cause of Death Statistics from 1996 to 2006 reported that the cardiovascular disease would become the number one cause of death in the next years. Therefore we realize that more accurate examination is required. The aim of this study was to investigate the accuracy of Calcium-scoring CT and the relationship between risk factor and quantitative scores of Calcium-scoring CT. Through this study we expect that the national public health will be improved. Seventy patients with chest pain were chosen at random. The patients were undergone both coronary CT antigraphy and Calcium - scoring CT at G hospital in Incheon from February 1 to June 30, 2008. The result of the Calcium-scoring CT showed its usefulness for Ischemic cardiovascular disease, with an accuracy similar to that of exercise/pharmacologic stress or ECG when it is difficult for a patient to exercise due to joint problems, aging or for other reasons.

  18. Improved attenuation correction for respiratory gated PET/CT with extended-duration cine CT: a simulation study

    Science.gov (United States)

    Zhang, Ruoqiao; Alessio, Adam M.; Pierce, Larry A.; Byrd, Darrin W.; Lee, Tzu-Cheng; De Man, Bruno; Kinahan, Paul E.

    2017-03-01

    Due to the wide variability of intra-patient respiratory motion patterns, traditional short-duration cine CT used in respiratory gated PET/CT may be insufficient to match the PET scan data, resulting in suboptimal attenuation correction that eventually compromises the PET quantitative accuracy. Thus, extending the duration of cine CT can be beneficial to address this data mismatch issue. In this work, we propose to use a long-duration cine CT for respiratory gated PET/CT, whose cine acquisition time is ten times longer than a traditional short-duration cine CT. We compare the proposed long-duration cine CT with the traditional short-duration cine CT through numerous phantom simulations with 11 respiratory traces measured during patient PET/CT scans. Experimental results show that, the long-duration cine CT reduces the motion mismatch between PET and CT by 41% and improves the overall reconstruction accuracy by 42% on average, as compared to the traditional short-duration cine CT. The long-duration cine CT also reduces artifacts in PET images caused by misalignment and mismatch between adjacent slices in phase-gated CT images. The improvement in motion matching between PET and CT by extending the cine duration depends on the patient, with potentially greater benefits for patients with irregular breathing patterns or larger diaphragm movements.

  19. Diagnostic accuracy of 64 multislice CT angiography in the assessment of coronary in-stent restenosis: A meta-analysis

    International Nuclear Information System (INIS)

    Sun Zhonghua; Almutairi, Abdulrahman Marzouq D.

    2010-01-01

    Purpose: The aim of this study was to perform a meta-analysis of the diagnostic accuracy of 64-slice CT angiography for the detection of coronary in-stent restenosis in patients treated with coronary stents when compared to conventional coronary angiography. Materials and methods: A search of PUBMED/MEDLINE, ProQuest and Cochrane library databases for English literature was performed. Only studies comparing 64-slice CT angiography with conventional coronary angiography for the detection of coronary in-stent restenosis (more than 50% stenosis) were included for analysis. Sensitivity and specificity estimates pooled across studies were tested using a fixed effects model. Results: Fourteen studies met selection criteria for inclusion in the analysis. The mean value of assessable stents was 89%. Prevalence of in-stent restenosis following coronary stenting was 20% among these studies. Pooled estimates of the sensitivity and specificity of overall 64-slice CT angiography for the detection of coronary in-stent restenosis was 90% (95% CI: 86%, 94%) and 91% (95% CI: 90%, 93%), respectively, based on the evaluation of assessable stents. Diagnostic value of 64-slice CT angiography was found to decrease significantly when the analysis was performed with inclusion of nonassessable segments in five studies, with pooled sensitivity and specificity being 79% (95% CI: 68%, 88%) and 81% (95% CI: 77%, 84%). Stent diameter is the main factor affecting the diagnostic value of 64-slice CT angiography. Conclusion: Our results showed that 64-slice CT angiography has high diagnostic value (both sensitivity and specificity) for detection of coronary in-stent restenosis based on assessable segments when compared to conventional coronary angiography.

  20. Diagnostic accuracy of low-dose versus ultra-low-dose CT for lumbar disc disease and facet joint osteoarthritis in patients with low back pain with MRI correlation

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Sun Hwa [Inje University College of Medicine, Department of Emergency Medicine, Sanggye Paik Hospital, Nowon-gu (Korea, Republic of); Yun, Seong Jong; Jo, Hyeon Hwan; Kim, Dong Hyeon [Republic of Korea Air Force, Department of Radiology, Aerospace Medical Center, Cheongwon-gun, Chungcheongbuk-do (Korea, Republic of); Song, Jae Gwang [Republic of Korea Air Force, Department of Orthopedic Surgery, Aerospace Medical Center, Cheongwon-gun, Chungcheongbuk-do (Korea, Republic of); Park, Yong Sung [Kyung Hee University Hospital at Gangdong, Kyung Hee University School of Medicine, Department of Radiology, Seoul (Korea, Republic of)

    2018-04-15

    To compare the image quality, radiation dose, and diagnostic performance between low-dose (LD) and ultra-low-dose (ULD) lumbar-spine (L-spine) CT with iterative reconstruction (IR) for patients with chronic low back pain (LBP). In total, 260 patients with chronic LBP who underwent L-spine CT between November 2015 and September 2016 were prospectively enrolled. Of these, 143 underwent LD-CT with IR and 117 underwent ULD-CT with IR. The patients were divided according to their body mass index (BMI) into BMI1 (<22.9 kg/m{sup 2}), BMI2 (23.0-24.9 kg/m{sup 2}), and BMI3 (≥25 kg/m{sup 2}) groups. Two blinded radiologists independently evaluated the signal-to-noise ratio (SNR), qualitative image quality, and final diagnoses (lumbar disc disease and facet joint osteoarthritis). L-spine MRIs interpreted by consensus were used as the reference standard. All data were statistically analyzed. ULD protocol showed significantly lower SNR for all patients (p < 0.001) except the vertebral bodies and lower qualitative image quality for BMI3 patients (p ≤ 0.033). There was no statistically significant difference between ULD (sensitivity, 95.1-98.1%; specificity, 92.5-98.7%; accuracy, 94.6-98.0%) and LD protocols (sensitivity, 95.6-100%; specificity, 95.5-98.9%; accuracy, 97.4-98.1%), (all p≥0.1) in the BMI1 and BMI2; while dose was 60-68% lower with the ULD protocol. Interobserver agreements were excellent or good with regard to image quality and final diagnoses. For the BM1 and BMI2 groups, ULD-CT provided an acceptable image quality and exhibited a diagnostic accuracy similar to that of LD-CT. These findings suggest that it is a useful diagnostic tool for patients with chronic LBP who exhibit a BMI of <25 kg/m{sup 2}. (orig.)

  1. Diagnostic accuracy of low-dose versus ultra-low-dose CT for lumbar disc disease and facet joint osteoarthritis in patients with low back pain with MRI correlation

    International Nuclear Information System (INIS)

    Lee, Sun Hwa; Yun, Seong Jong; Jo, Hyeon Hwan; Kim, Dong Hyeon; Song, Jae Gwang; Park, Yong Sung

    2018-01-01

    To compare the image quality, radiation dose, and diagnostic performance between low-dose (LD) and ultra-low-dose (ULD) lumbar-spine (L-spine) CT with iterative reconstruction (IR) for patients with chronic low back pain (LBP). In total, 260 patients with chronic LBP who underwent L-spine CT between November 2015 and September 2016 were prospectively enrolled. Of these, 143 underwent LD-CT with IR and 117 underwent ULD-CT with IR. The patients were divided according to their body mass index (BMI) into BMI1 (<22.9 kg/m 2 ), BMI2 (23.0-24.9 kg/m 2 ), and BMI3 (≥25 kg/m 2 ) groups. Two blinded radiologists independently evaluated the signal-to-noise ratio (SNR), qualitative image quality, and final diagnoses (lumbar disc disease and facet joint osteoarthritis). L-spine MRIs interpreted by consensus were used as the reference standard. All data were statistically analyzed. ULD protocol showed significantly lower SNR for all patients (p < 0.001) except the vertebral bodies and lower qualitative image quality for BMI3 patients (p ≤ 0.033). There was no statistically significant difference between ULD (sensitivity, 95.1-98.1%; specificity, 92.5-98.7%; accuracy, 94.6-98.0%) and LD protocols (sensitivity, 95.6-100%; specificity, 95.5-98.9%; accuracy, 97.4-98.1%), (all p≥0.1) in the BMI1 and BMI2; while dose was 60-68% lower with the ULD protocol. Interobserver agreements were excellent or good with regard to image quality and final diagnoses. For the BM1 and BMI2 groups, ULD-CT provided an acceptable image quality and exhibited a diagnostic accuracy similar to that of LD-CT. These findings suggest that it is a useful diagnostic tool for patients with chronic LBP who exhibit a BMI of <25 kg/m 2 . (orig.)

  2. Assessment of CT numbers in limited and medium field-of-view scans taken using Accuitomo 170 and Veraviewepocs 3De cone-beam computed tomography scanners

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Matheus L. [Dept. of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, Campinas (Brazil); Tosoni, Guilherme M. [Dept. of Oral Diagnosis and Surgery, Araraquara Dental School, Sao Paulo State University, Araraquara (Brazil); Lindsey, David H.; Mendoza, Kristopher; Tetradis, Sotirios; Mallya, Sanjay M. [Section of Oral and Maxillofacial Radiology, School of Dentistry, University of California, Los Angeles (United States)

    2014-12-15

    To assess the influence of anatomic location on the relationship between computed tomography (CT) number and X-ray attenuation in limited and medium field-of-view (FOV) scans. Tubes containing solutions with different concentrations of K2HPO4 were placed in the tooth sockets of a human head phantom. Cone-beam computed tomography (CBCT) scans were acquired, and CT numbers of the K{sub 2}HPO{sub 4} solutions were measured. The relationship between CT number and K{sub 2}HPO{sub 4} concentration was examined by linear regression analyses. Then, the variation in CT number according to anatomic location was examined. The relationship between K{sub 2}HPO{sub 4} concentration and CT number was strongly linear. The slopes of the linear regressions for the limited FOVs were almost 2-fold lower than those for the medium FOVs. The absolute CT number differed between imaging protocols and anatomic locations. There is a strong linear relationship between X-ray attenuation and CT number. The specific imaging protocol and anatomic location of the object strongly influence this relationship.

  3. In-vivo and ex-vivo assessment of the accuracy of the computer-aided volumetry of porcine kidney in CT images

    Energy Technology Data Exchange (ETDEWEB)

    Cai, W.; Harris, G.; Holalkere, N.; Sahani, D.; Yoshida, H. [Massachusetts General Hospital and Harvard Medical School (United States)

    2007-06-15

    Measurement of kidney volume by computed tomography (CT), called renal volumetry, is indispensable for diagnosis and treatment of kidney-related diseases. Computer-aided volumetry (CAV) of kidney relies on an efficient and accurate segmentation method of the kidney. The purpose of this study is to assess the accuracy of our CAV of kidney scheme using dynamic-threshold (DT) level set method, based on in-vivo and ex-vivo reference standards. Eight Yorkshire breed anesthetized pigs were scanned on a 64-slice multi-detector CT scanner (Sensation 64, Siemens) after an injection of iodinated (300 mgl/ml) contrast agent through an IV cannula. The kidneys of the pigs were then surgically resected and scanned on CT in the same manner. Both in-vivo and ex-vivo CT images were subjected to our volumetry scheme. The resulting volumes of the kidneys were compared with the in-vivo and ex-vivo reference standards: the former was established by manual contouring of the kidneys on the CT images by an experienced radiologist, and the latter was established as the water displacement volume of the resected kidney. Our CAV of kidney scheme demonstrated accurate in-vivo and ex-vivo measurement of kidney volume, despite a large difference between the two reference standards. (orig.)

  4. In-vivo and ex-vivo assessment of the accuracy of the computer-aided volumetry of porcine kidney in CT images

    International Nuclear Information System (INIS)

    Cai, W.; Harris, G.; Holalkere, N.; Sahani, D.; Yoshida, H.

    2007-01-01

    Measurement of kidney volume by computed tomography (CT), called renal volumetry, is indispensable for diagnosis and treatment of kidney-related diseases. Computer-aided volumetry (CAV) of kidney relies on an efficient and accurate segmentation method of the kidney. The purpose of this study is to assess the accuracy of our CAV of kidney scheme using dynamic-threshold (DT) level set method, based on in-vivo and ex-vivo reference standards. Eight Yorkshire breed anesthetized pigs were scanned on a 64-slice multi-detector CT scanner (Sensation 64, Siemens) after an injection of iodinated (300 mgl/ml) contrast agent through an IV cannula. The kidneys of the pigs were then surgically resected and scanned on CT in the same manner. Both in-vivo and ex-vivo CT images were subjected to our volumetry scheme. The resulting volumes of the kidneys were compared with the in-vivo and ex-vivo reference standards: the former was established by manual contouring of the kidneys on the CT images by an experienced radiologist, and the latter was established as the water displacement volume of the resected kidney. Our CAV of kidney scheme demonstrated accurate in-vivo and ex-vivo measurement of kidney volume, despite a large difference between the two reference standards. (orig.)

  5. Detection of pulmonary metastatic nodules: usefulness of low-dose multidetector CT in comparison with chest radiograph

    International Nuclear Information System (INIS)

    Kim, Ki Nam; Lee, Ki Nam; Yang, Doo Kyung; Lee, Soo Keol

    2006-01-01

    We wanted to evaluate the usefulness of low-dose multidetector CT for the detection and follow-up of pulmonary metastatic nodules in patients suffering with malignancy. We retrospectively reviewed the conventional chest radiographs and low-dose multidetector CT (low-dose CT) scans of 81 patients who had been under the diagnosis of malignancy. We reviewed the detection of pulmonary nodules and we counted the number of nodules detected by each method. The nodules were confirmed by surgical operation and by the radiologic criteria. The accuracy, sensitivity, specificity and positive and negative predictive values of each method for detecting metastatic nodules were compared with χ 2 tests. Low-dose CT depicted more nodules than did chest radiograph, and the indeterminate nodules seen on chest radiograph may be clearly benign on low-dose CT (eg. calcified granulomas or bony lesions). The accuracy of low-dose CT (75.3%) was significantly higher than that of chest radiograph (49.4%) for the detection for metastatic nodules (ρ < 0.05). Low-dose CT may provide better information than does chest radiograph for diagnosing pulmonary metastasis

  6. Additional value of PET-CT in the staging of lung cancer: comparison with CT alone, PET alone and visual correlation of PET and CT

    International Nuclear Information System (INIS)

    Wever, W. de; Marchal, G.; Bogaert, J.; Verschakelen, J.A.; Ceyssens, S.; Mortelmans, L.; Stroobants, S.

    2007-01-01

    Integrated positron emission tomography (PET) and computed tomography (CT) is a new imaging modality offering anatomic and metabolic information. The purpose was to evaluate retrospectively the accuracy of integrated PET-CT in the staging of a suggestive lung lesion, comparing this with the accuracy of CT alone, PET alone and visually correlated PET-CT. Fifty patients undergoing integrated PET-CT for staging of a suggestive lung lesion were studied. Their tumor, node, metastasis (TNM) statuses were determined with CT, PET, visually correlated PET-CT and integrated PET-CT. These TNM stages were compared with the surgical TNM status. Integrated PET-CT was the most accurate imaging technique in the assessment of the TNM status. Integrated PET-CT predicted correctly the T status, N status, M status and TNM status in, respectively, 86%, 80%, 98%, 70% versus 68%, 66%,88%, 46% with CT, 46%, 70%, 96%, 30% with PET and 72%, 68%, 96%, 54% with visually correlated PET-CT. T status and N status were overstaged, respectively, in 8% and 16% with integrated PET-CT, in 20% and 28% with CT, in 16% and 20% with PET, in 12% and 20% with visually correlated PET-CT and understaged in 6% and 4% with integrated PET-CT, versus 12% and 6% with CT, 38% and 10% with PET and 12% with visually correlated PET-CT. Integrated PET-CT improves the staging of lung cancer through a better anatomic localization and characterization of lesions and is superior to CT alone and PET alone. If this technique is not available, visual correlation of PET and CT can be a valuable alternative. (orig.)

  7. Diagnosis of clinical staging of bladder cancer by CT and angiography

    International Nuclear Information System (INIS)

    Kobayashi, Isao; Igawa, Mikio; Ohnishi, Yoshio; Nakano, Hiroshi; Nihira, Hiromi; Mori, Masaki; Okada, Mitsuo.

    1984-01-01

    The preoperative staging of bladder cancer is of fundamental importance for prognostic evaluation and surgical indication. We studied the accuracy of computed tomography (CT) and angiography in defining the extent of local invasion in 16 patients with surgically proven carcinoma of the bladder. The overall accuracy of CT and angiographic staging in these cases was 75 % and 50 % respectively. In low stage, the accuracy was 90 % in CT and 70 % in angiography. In high stage, the accuracy was 50 % in CT and 16.7 % in angiography. Our results seems to indicate lower accuracy in high stage bladder cancer compared with other research. Data from a much larger series are required to ascertain whether the additional information provided by CT and angiography will produce any improvement in patient management. (author)

  8. Intervertebral anticollision constraints improve out-of-plane translation accuracy of a single-plane fluoroscopy-to-CT registration method for measuring spinal motion

    International Nuclear Information System (INIS)

    Lin, Cheng-Chung; Tsai, Tsung-Yuan; Hsu, Shih-Jung; Lu, Tung-Wu; Shih, Ting-Fang; Wang, Ting-Ming

    2013-01-01

    Purpose: The study aimed to propose a new single-plane fluoroscopy-to-CT registration method integrated with intervertebral anticollision constraints for measuring three-dimensional (3D) intervertebral kinematics of the spine; and to evaluate the performance of the method without anticollision and with three variations of the anticollision constraints via an in vitro experiment. Methods: The proposed fluoroscopy-to-CT registration approach, called the weighted edge-matching with anticollision (WEMAC) method, was based on the integration of geometrical anticollision constraints for adjacent vertebrae and the weighted edge-matching score (WEMS) method that matched the digitally reconstructed radiographs of the CT models of the vertebrae and the measured single-plane fluoroscopy images. Three variations of the anticollision constraints, namely, T-DOF, R-DOF, and A-DOF methods, were proposed. An in vitro experiment using four porcine cervical spines in different postures was performed to evaluate the performance of the WEMS and the WEMAC methods. Results: The WEMS method gave high precision and small bias in all components for both vertebral pose and intervertebral pose measurements, except for relatively large errors for the out-of-plane translation component. The WEMAC method successfully reduced the out-of-plane translation errors for intervertebral kinematic measurements while keeping the measurement accuracies for the other five degrees of freedom (DOF) more or less unaltered. The means (standard deviations) of the out-of-plane translational errors were less than −0.5 (0.6) and −0.3 (0.8) mm for the T-DOF method and the R-DOF method, respectively. Conclusions: The proposed single-plane fluoroscopy-to-CT registration method reduced the out-of-plane translation errors for intervertebral kinematic measurements while keeping the measurement accuracies for the other five DOF more or less unaltered. With the submillimeter and subdegree accuracy, the WEMAC method was

  9. Determination of CT number and density profile of binderless, pre-treated and tannin-based Rhizophora spp. particleboards using computed tomography imaging and electron density phantom

    Energy Technology Data Exchange (ETDEWEB)

    Yusof, Mohd Fahmi Mohd, E-mail: mfahmi@usm.my; Hamid, Puteri Nor Khatijah Abdul; Tajuddin, Abdul Aziz [School of Physics, Universiti Sains Malaysia, 11800 Penang (Malaysia); Bauk, Sabar [School of Distance Education, Universiti Sains Malaysia, 11800 Penang (Malaysia); Hashim, Rokiah [School of Industrial Technologies, Universiti Sains Malaysia, 11800 Penang (Malaysia)

    2015-04-29

    Plug density phantoms were constructed in accordance to CT density phantom model 062M CIRS using binderless, pre-treated and tannin-based Rhizophora Spp. particleboards. The Rhizophora Spp. plug phantoms were scanned along with the CT density phantom using Siemens Somatom Definition AS CT scanner at three CT energies of 80, 120 and 140 kVp. 15 slices of images with 1.0 mm thickness each were taken from the central axis of CT density phantom for CT number and CT density profile analysis. The values were compared to water substitute plug phantom from the CT density phantom. The tannin-based Rhizophora Spp. gave the nearest value of CT number to water substitute at 80 and 120 kVp CT energies with χ{sup 2} value of 0.011 and 0.014 respectively while the binderless Rhizphora Spp. gave the nearest CT number to water substitute at 140 kVp CT energy with χ{sup 2} value of 0.023. The tannin-based Rhizophora Spp. gave the nearest CT density profile to water substitute at all CT energies. This study indicated the suitability of Rhizophora Spp. particleboard as phantom material for the use in CT imaging studies.

  10. Determination of CT number and density profile of binderless, pre-treated and tannin-based Rhizophora spp. particleboards using computed tomography imaging and electron density phantom

    International Nuclear Information System (INIS)

    Yusof, Mohd Fahmi Mohd; Hamid, Puteri Nor Khatijah Abdul; Tajuddin, Abdul Aziz; Bauk, Sabar; Hashim, Rokiah

    2015-01-01

    Plug density phantoms were constructed in accordance to CT density phantom model 062M CIRS using binderless, pre-treated and tannin-based Rhizophora Spp. particleboards. The Rhizophora Spp. plug phantoms were scanned along with the CT density phantom using Siemens Somatom Definition AS CT scanner at three CT energies of 80, 120 and 140 kVp. 15 slices of images with 1.0 mm thickness each were taken from the central axis of CT density phantom for CT number and CT density profile analysis. The values were compared to water substitute plug phantom from the CT density phantom. The tannin-based Rhizophora Spp. gave the nearest value of CT number to water substitute at 80 and 120 kVp CT energies with χ 2 value of 0.011 and 0.014 respectively while the binderless Rhizphora Spp. gave the nearest CT number to water substitute at 140 kVp CT energy with χ 2 value of 0.023. The tannin-based Rhizophora Spp. gave the nearest CT density profile to water substitute at all CT energies. This study indicated the suitability of Rhizophora Spp. particleboard as phantom material for the use in CT imaging studies

  11. Design and validation of a CT-guided robotic system for lung cancer brachytherapy.

    Science.gov (United States)

    Dou, Huaisu; Jiang, Shan; Yang, Zhiyong; Sun, Luqing; Ma, Xiaodong; Huo, Bin

    2017-09-01

    of the robotic system was 0.49 ± 0.29 mm. Phantom experiments indicated that the accuracy of needle insertion was 1.5 ± 1.7 mm at a depth ranging from 30 to 80 mm. The time used to adjust the template to the target position was 12 min on average by robotic system automatically. An average of 30 min was saved compared with the manual positioning procedure in phantom experiments. This paper describes the design and experimental validation of a novel CT-guided robotic system for lung cancer brachytherapy. The robotic system was validated in a number of aspects which prove that it was capable of locating the template with clinically acceptable accuracy in the CT environment. All experimental results indicated that the system is reliable and ready to be applied to further studies on animals. © 2017 American Association of Physicists in Medicine.

  12. Poster — Thur Eve — 11: Validation of the orthopedic metallic artifact reduction tool for CT simulations at the Ottawa Hospital Cancer Centre

    International Nuclear Information System (INIS)

    Sutherland, J; Foottit, C

    2014-01-01

    Metallic implants in patients can produce image artifacts in kilovoltage CT simulation images which can introduce noise and inaccuracies in CT number, affecting anatomical segmentation and dose distributions. The commercial orthopedic metal artifact reduction algorithm (O-MAR) (Philips Healthcare System) was recently made available on CT simulation scanners at our institution. This study validated the clinical use of O-MAR by investigating its effects on CT number and dose distributions. O-MAR corrected and uncorrected images were acquired with a Philips Brilliance Big Bore CT simulator of a cylindrical solid water phantom that contained various plugs (including metal) of known density. CT number accuracy was investigated by determining the mean and standard deviation in regions of interest (ROI) within each plug for uncorrected and O-MAR corrected images and comparing with no-metal image values. Dose distributions were calculated using the Monaco treatment planning system. Seven open fields were equally spaced about the phantom around a ROI near the center of the phantom. These were compared to a “correct” dose distribution calculated by overriding electron densities a no-metal phantom image to produce an image containing metal but no artifacts. An overall improvement in CT number and dose distribution accuracy was achieved by applying the O-MAR correction. Mean CT numbers and standard deviations were found to be generally improved. Exceptions included lung equivalent media, which is consistent with vendor specified contraindications. Dose profiles were found to vary by ±4% between uncorrected or O-MAR corrected images with O-MAR producing doses closer to ground truth

  13. Poster — Thur Eve — 11: Validation of the orthopedic metallic artifact reduction tool for CT simulations at the Ottawa Hospital Cancer Centre

    Energy Technology Data Exchange (ETDEWEB)

    Sutherland, J; Foottit, C [The Ottawa Hospital Cancer Centre (Canada)

    2014-08-15

    Metallic implants in patients can produce image artifacts in kilovoltage CT simulation images which can introduce noise and inaccuracies in CT number, affecting anatomical segmentation and dose distributions. The commercial orthopedic metal artifact reduction algorithm (O-MAR) (Philips Healthcare System) was recently made available on CT simulation scanners at our institution. This study validated the clinical use of O-MAR by investigating its effects on CT number and dose distributions. O-MAR corrected and uncorrected images were acquired with a Philips Brilliance Big Bore CT simulator of a cylindrical solid water phantom that contained various plugs (including metal) of known density. CT number accuracy was investigated by determining the mean and standard deviation in regions of interest (ROI) within each plug for uncorrected and O-MAR corrected images and comparing with no-metal image values. Dose distributions were calculated using the Monaco treatment planning system. Seven open fields were equally spaced about the phantom around a ROI near the center of the phantom. These were compared to a “correct” dose distribution calculated by overriding electron densities a no-metal phantom image to produce an image containing metal but no artifacts. An overall improvement in CT number and dose distribution accuracy was achieved by applying the O-MAR correction. Mean CT numbers and standard deviations were found to be generally improved. Exceptions included lung equivalent media, which is consistent with vendor specified contraindications. Dose profiles were found to vary by ±4% between uncorrected or O-MAR corrected images with O-MAR producing doses closer to ground truth.

  14. CT coronary angiography: Influence of different cardiac reconstruction intervals on image quality and diagnostic accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Dewey, Marc [Department of Radiology, Charite Medical School, Humboldt-Universitaet zu Berlin (Germany)], E-mail: marc.dewey@charite.de; Teige, Florian [Department of Radiology, Charite Medical School, Humboldt-Universitaet zu Berlin (Germany); Rutsch, Wolfgang [Department of Cardiology, Charite Medical School, Humboldt-Universitaet zu Berlin (Germany)], E-mail: wolfgang.rutsch@charite.de; Schink, Tania [Department of Medical Biometry, Charite Medical School, Humboldt-Universitaet zu Berlin (Germany)], E-mail: peter.martus@charite.de; Hamm, Bernd [Department of Radiology, Charite Medical School, Humboldt-Universitaet zu Berlin (Germany)

    2008-07-15

    Purpose: To prospectively analyze image quality and diagnostic accuracy of different reconstruction intervals of coronary angiography using multislice computed tomography (MSCT). Materials and methods: For each of 47 patients, 10 ECG-gated MSCT reconstructions were generated throughout the RR interval from 0 to 90%, resulting in altogether 470 datasets. These datasets were randomly analyzed for image quality and accuracy and compared with conventional angiography. Statistical comparison of intervals was performed using nonparametric analysis for repeated measurements to account for clustering of arteries within patients. Results: Image reconstruction intervals centered at 80, 70, and 40% of the RR interval resulted (in that order) in the best overall image quality for all four main coronary vessels. Eighty percent reconstructions also yielded the highest diagnostic accuracy of all intervals. The combination of the three best intervals (80, 70, and 40%) significantly reduced the nondiagnostic rate as compared with 80% alone (p = 0.005). However, the optimal reconstruction interval combination achieved significantly improved specificities and nondiagnostic rates (p < 0.05). The optimal combination consisted of 1.7 {+-} 0.9 reconstruction intervals on average. In approximately half of the patients (49%, 23/47) a single reconstruction was optimal. In 18 (38%), 3 (6%), and 3 (6%) patients one, two, and three additional reconstruction intervals were required, respectively, to achieve optimal quality. In 28% of the patients the optimal combination consisted of reconstructions other than the three best intervals (80, 70, and 40%). Conclusion: Multiple image reconstruction intervals are essential to ensure high image quality and accuracy of CT coronary angiography.

  15. Effect of CT image size and resolution on the accuracy of rock property estimates

    Science.gov (United States)

    Bazaikin, Y.; Gurevich, B.; Iglauer, S.; Khachkova, T.; Kolyukhin, D.; Lebedev, M.; Lisitsa, V.; Reshetova, G.

    2017-05-01

    In order to study the effect of the micro-CT scan resolution and size on the accuracy of upscaled digital rock property estimation of core samples Bentheimer sandstone images with the resolution varying from 0.9 μm to 24 μm are used. We statistically show that the correlation length of the pore-to-matrix distribution can be reliably determined for the images with the resolution finer than 9 voxels per correlation length and the representative volume for this property is about 153 correlation length. Similar resolution values for the statistically representative volume are also valid for the estimation of the total porosity, specific surface area, mean curvature, and topology of the pore space. Only the total porosity and the number of isolated pores are stably recovered, whereas geometry and the topological measures of the pore space are strongly affected by the resolution change. We also simulate fluid flow in the pore space and estimate permeability and tortuosity of the sample. The results demonstrate that the representative volume for the transport property calculation should be greater than 50 correlation lengths of pore-to-matrix distribution. On the other hand, permeability estimation based on the statistical analysis of equivalent realizations shows some weak influence of the resolution on the transport properties. The reason for this might be that the characteristic scale of the particular physical processes may affect the result stronger than the model (image) scale.

  16. Accuracy of automated software-guided detection of significant coronary artery stenosis by CT angiography: comparison with invasive catheterisation

    International Nuclear Information System (INIS)

    Anders, Katharina; Uder, Michael; Achenbach, Stephan; Petit, Isabel; Daniel, Werner G.; Pflederer, Tobias

    2013-01-01

    True automated detection of coronary artery stenoses might be useful whenever expert evaluation is not available, or as a ''second reader'' to enhance diagnostic confidence. We evaluated the accuracy of a PC-based stenosis detection tool alone and combined with expert interpretation. One hundred coronary CT angiography datasets were evaluated with the automated software alone, by manual interpretation (axial images, multiplanar reformations and maximum intensity projections in free double-oblique planes), and by expert interpretation aware of the automated findings. Stenoses ≥ 50 % were noted per-vessel and per-patient, and compared with invasive angiography. Automated post-processing was successful in 90 % of patients (88 % of vessels). When excluding uninterpretable datasets, per-patient sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) were 89 %, 79 %, 74 % and 92 % (per-vessel: 82 %, 85 %, 48 % and 96 %). All 100 datasets were evaluable by expert interpretation. Per-patient sensitivity, specificity, PPV and NPV were 95 %, 95 %, 93 % and 97 % (per-vessel: 89 %,98 %, 88 % and 98 %). Knowing the results of automated interpretation did not improve the performance of expert readers. Automated off-line post-processing of coronary CT angiography shows adequate sensitivity, but relatively low specificity in coronary stenosis detection. It does not increase accuracy of expert interpretation. Failure of post-processing in 10 % of all patients necessitates additional manual image work-up. (orig.)

  17. Accuracy of Combined Computed Tomography Colonography and Dual Energy Iiodine Map Imaging for Detecting Colorectal masses using High-pitch Dual-source CT.

    Science.gov (United States)

    Sun, Kai; Han, Ruijuan; Han, Yang; Shi, Xuesen; Hu, Jiang; Lu, Bin

    2018-02-28

    To evaluate the diagnostic accuracy of combined computed tomography colonography (CTC) and dual-energy iodine map imaging for detecting colorectal masses using high-pitch dual-source CT, compared with optical colonography (OC) and histopathologic findings. Twenty-eight consecutive patients were prospectively enrolled in this study. All patients were underwent contrast-enhanced CTC acquisition using dual-energy mode and OC and pathologic examination. The size of the space-occupied mass, the CT value after contrast enhancement, and the iodine value were measured and statistically compared. The sensitivity, specificity, accuracy rate, and positive predictive and negative predictive values of dual-energy contrast-enhanced CTC were calculated and compared between conventional CTC and dual-energy iodine images. The iodine value of stool was significantly lower than the colonic neoplasia (P dual-energy iodine maps imaging was 95.6% (95% CI = 77.9%-99.2%). The specificity of the two methods was 42.8% (95% CI = 15.4%-93.5%) and 100% (95% CI = 47.9%-100%; P = 0.02), respectively. Compared with optical colonography and histopathology, combined CTC and dual-energy iodine maps imaging can distinguish stool and colonic neoplasia, distinguish between benign and malignant tumors initially and improve the diagnostic accuracy of CTC for colorectal cancer screening.

  18. Comparison of CT Fluoroscopy-Guided Manual and CT-Guided Robotic Positioning System for In Vivo Needle Placements in Swine Liver

    International Nuclear Information System (INIS)

    Cornelis, F.; Takaki, H.; Laskhmanan, M.; Durack, J. C.; Erinjeri, J. P.; Getrajdman, G. I.; Maybody, M.; Sofocleous, C. T.; Solomon, S. B.; Srimathveeravalli, G.

    2015-01-01

    PurposeTo compare CT fluoroscopy-guided manual and CT-guided robotic positioning system (RPS)-assisted needle placement by experienced IR physicians to targets in swine liver.Materials and MethodsManual and RPS-assisted needle placement was performed by six experienced IR physicians to four 5 mm fiducial seeds placed in swine liver (n = 6). Placement performance was assessed for placement accuracy, procedure time, number of confirmatory scans, needle manipulations, and procedure radiation dose. Intra-modality difference in performance for each physician was assessed using paired t test. Inter-physician performance variation for each modality was analyzed using Kruskal–Wallis test.ResultsPaired comparison of manual and RPS-assisted placements to a target by the same physician indicated accuracy outcomes was not statistically different (manual: 4.53 mm; RPS: 4.66 mm; p = 0.41), but manual placement resulted in higher total radiation dose (manual: 1075.77 mGy/cm; RPS: 636.4 mGy/cm; p = 0.03), required more confirmation scans (manual: 6.6; RPS: 1.6; p < 0.0001) and needle manipulations (manual: 4.6; RPS: 0.4; p < 0.0001). Procedure time for RPS was longer than manual placement (manual: 6.12 min; RPS: 9.7 min; p = 0.0003). Comparison of inter-physician performance during manual placement indicated significant differences in the time taken to complete placements (p = 0.008) and number of repositions (p = 0.04) but not in other study measures (p > 0.05). Comparison of inter-physician performance during RPS-assisted placement suggested statistically significant differences in procedure time (p = 0.02) and not in other study measures (p > 0.05).ConclusionsCT-guided RPS-assisted needle placement reduced radiation dose, number of confirmatory scans, and needle manipulations when compared to manual needle placement by experienced IR physicians, with equivalent accuracy

  19. Comparison of CT Fluoroscopy-Guided Manual and CT-Guided Robotic Positioning System for In Vivo Needle Placements in Swine Liver

    Energy Technology Data Exchange (ETDEWEB)

    Cornelis, F.; Takaki, H. [Memorial Sloan-Kettering Cancer Center, Interventional Radiology Service, Department of Radiology (United States); Laskhmanan, M. [Perfint Healthcare Inc (India); Durack, J. C.; Erinjeri, J. P.; Getrajdman, G. I.; Maybody, M.; Sofocleous, C. T.; Solomon, S. B.; Srimathveeravalli, G., E-mail: srimaths@mskcc.org [Memorial Sloan-Kettering Cancer Center, Interventional Radiology Service, Department of Radiology (United States)

    2015-10-15

    PurposeTo compare CT fluoroscopy-guided manual and CT-guided robotic positioning system (RPS)-assisted needle placement by experienced IR physicians to targets in swine liver.Materials and MethodsManual and RPS-assisted needle placement was performed by six experienced IR physicians to four 5 mm fiducial seeds placed in swine liver (n = 6). Placement performance was assessed for placement accuracy, procedure time, number of confirmatory scans, needle manipulations, and procedure radiation dose. Intra-modality difference in performance for each physician was assessed using paired t test. Inter-physician performance variation for each modality was analyzed using Kruskal–Wallis test.ResultsPaired comparison of manual and RPS-assisted placements to a target by the same physician indicated accuracy outcomes was not statistically different (manual: 4.53 mm; RPS: 4.66 mm; p = 0.41), but manual placement resulted in higher total radiation dose (manual: 1075.77 mGy/cm; RPS: 636.4 mGy/cm; p = 0.03), required more confirmation scans (manual: 6.6; RPS: 1.6; p < 0.0001) and needle manipulations (manual: 4.6; RPS: 0.4; p < 0.0001). Procedure time for RPS was longer than manual placement (manual: 6.12 min; RPS: 9.7 min; p = 0.0003). Comparison of inter-physician performance during manual placement indicated significant differences in the time taken to complete placements (p = 0.008) and number of repositions (p = 0.04) but not in other study measures (p > 0.05). Comparison of inter-physician performance during RPS-assisted placement suggested statistically significant differences in procedure time (p = 0.02) and not in other study measures (p > 0.05).ConclusionsCT-guided RPS-assisted needle placement reduced radiation dose, number of confirmatory scans, and needle manipulations when compared to manual needle placement by experienced IR physicians, with equivalent accuracy.

  20. Measurement of anatomical structure of jaw bone trabecula with micro-CT and its accuracy evaluation

    International Nuclear Information System (INIS)

    Jin Guangchun; Zhang Xiaoyan; Jiang Ling; Li Xianglin; Zhang Di; Li Weixing; Jin Xizhen; Jin Dongchun

    2011-01-01

    Objective: To measure the anatomic structures of the jaw and to discuss the accuracy of the method. Methods: The mandibular specimens were divided into ankle condylar group (3 specimens, regular trabecular architecture and mandibular body group (5 specimens, irregular trabecular trabecular architecture). Fifteen volumes of interests (VOI) were created in each group and the total of thirty were measured. The mean value and standard deviation of each parameter at each reconstruction voxel size were calculated. Repeated ANOVA test was used to determine whether the significant differences in the values existed between each parameter. Results: The structure model index (SMI) value in mandibular body group had significant difference (P<0.05) at all reconstruction voxel sizes compared with 18 μm group; but in condyle group, the reconstruction voxel size of 36 μm was not significant (P>0.05). The differences of trabecular thickness (Tb. Th), trabecular number (Tb. N) and trabecular separation (Tb. Sp) values between condyle and mandibular body groups were significant at all reconstruction voxel sizes (P<0.05). In condyle group, except for Tb. Th, Tb. N and Tb. Sp, the most parameter values were not significant at reconstruction pixel size of 36μm. In the mandible body group, the differences of all parameter values between different pixel sizes were significant. Conclusion: Micro-CT can reflect the anatomical changes of bone trabecula structure. (authors)

  1. CT-guided core needle biopsy of pleural lesions: Evaluating diagnostic yield and associated complications

    International Nuclear Information System (INIS)

    Niu, Xiang Ke; Bhetuwal, Anup; Yang, Han Feng

    2015-01-01

    The purpose of this study was to retrospectively evaluate the diagnostic accuracy and complications of CT-guided core needle biopsy (CT-guided CNB) of pleural lesion and the possible effects of influencing factors. From September 2007 to June 2013, 88 consecutive patients (60 men and 28 women; mean [+/- standard deviation] age, 51.1 +/- 14.4 years; range, 19-78 years) underwent CT-guided CNB, which was performed by two experienced chest radiologists in our medical center. Out of 88 cases, 56 (63%) were diagnosed as malignant, 28 (31%) as benign and 4 (5%) as indeterminate for CNB of pleural lesions. The final diagnosis was confirmed by either histopathological diagnosis or clinical follow-up. The diagnostic accuracy, sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and complication rates were statistically evaluated. Influencing factors (patient age, sex, lesion size, pleural-puncture angle, patient position, pleural effusion, and number of pleural punctures) were assessed for their effect on accuracy of CT-guided CNB using univariate and subsequent multivariate analysis. Diagnostic accuracy, sensitivity, specificity, PPV, and NPV were 89.2%, 86.1%, 100%, 100%, and 67.8%, respectively. The influencing factors had no significant effect in altering diagnostic accuracy. As far as complications were concerned, occurrence of pneumothorax was observed in 14 (16%) out of 88 patients. Multivariate analysis revealed lesion size/pleural thickening as a significant risk factor (odds ratio [OR]: 8.744, p = 0.005) for occurrence of pneumothorax. Moreover, presence of pleural effusion was noted as a significant protective factor (OR: 0.171, p = 0.037) for pneumothorax. CT-guided CNB of pleural lesion is a safe procedure with high diagnostic yield and low risk of significant complications.

  2. CT-guided core needle biopsy of pleural lesions: Evaluating diagnostic yield and associated complications

    Energy Technology Data Exchange (ETDEWEB)

    Niu, Xiang Ke [Dept. of Radiology, Affiliated Hospital of Chengdu University, Chengdu (China); Bhetuwal, Anup; Yang, Han Feng [Dept. of Radiology, Sichuan Key Laboratory of Medical Imaging, Affiliated Hospital of North Sichuan Medical College, Nanchong (China)

    2015-02-15

    The purpose of this study was to retrospectively evaluate the diagnostic accuracy and complications of CT-guided core needle biopsy (CT-guided CNB) of pleural lesion and the possible effects of influencing factors. From September 2007 to June 2013, 88 consecutive patients (60 men and 28 women; mean [+/- standard deviation] age, 51.1 +/- 14.4 years; range, 19-78 years) underwent CT-guided CNB, which was performed by two experienced chest radiologists in our medical center. Out of 88 cases, 56 (63%) were diagnosed as malignant, 28 (31%) as benign and 4 (5%) as indeterminate for CNB of pleural lesions. The final diagnosis was confirmed by either histopathological diagnosis or clinical follow-up. The diagnostic accuracy, sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and complication rates were statistically evaluated. Influencing factors (patient age, sex, lesion size, pleural-puncture angle, patient position, pleural effusion, and number of pleural punctures) were assessed for their effect on accuracy of CT-guided CNB using univariate and subsequent multivariate analysis. Diagnostic accuracy, sensitivity, specificity, PPV, and NPV were 89.2%, 86.1%, 100%, 100%, and 67.8%, respectively. The influencing factors had no significant effect in altering diagnostic accuracy. As far as complications were concerned, occurrence of pneumothorax was observed in 14 (16%) out of 88 patients. Multivariate analysis revealed lesion size/pleural thickening as a significant risk factor (odds ratio [OR]: 8.744, p = 0.005) for occurrence of pneumothorax. Moreover, presence of pleural effusion was noted as a significant protective factor (OR: 0.171, p = 0.037) for pneumothorax. CT-guided CNB of pleural lesion is a safe procedure with high diagnostic yield and low risk of significant complications.

  3. Prospective evaluation of MRI, 11C-acetate PET/CT and contrast-enhanced CT for staging of bladder cancer

    International Nuclear Information System (INIS)

    Vargas, H.A.; Akin, O.; Schöder, H.; Olgac, S.; Dalbagni, G.; Hricak, H.; Bochner, B.H.

    2012-01-01

    Purpose: To prospectively evaluate the diagnostic performance of magnetic resonance imaging (MRI), 11 C-acetate positron emission tomography/computed tomography (PET/CT) and contrast-enhanced CT for bladder cancer staging, using whole-mount pathologic review of radical cystectomy and pelvic lymph node specimens as the reference standard. Materials and methods: The institutional review board approved this prospective study, which was compliant with the Health Insurance Portability and Accountability Act. Written informed consent was obtained from 16 patients with histologically confirmed bladder cancer who underwent MRI, 11 C-acetate PET/CT and contrast-enhanced CT before radical cystectomy and pelvic lymph node dissection. Before imaging 4/16 patients had received intravesical Bacillus Calmette-Guérin treatment, 6 had received systemic chemotherapy, 3 had received both and 3 had received neither. Measures of diagnostic performance including accuracy, sensitivity and specificity were estimated separately for each imaging modality. Results: MRI correctly staged 56% of patients (9/16), overstaged 38% (6/16) and understaged 6% (1/16). CT correctly staged 50% of patients (8/16), overstaged 44% (7/16) and understaged 6% (1/16). In 9 patients, 11 C-acetate PET/CT showed uptake within the bladder wall; the uptake was true-positive in 7 patients and false-positive in 2 patients. Of the remaining 7 patients, 5 had true-negative and 2 had false-negative PET/CT results for cancer in the bladder wall. For all modalities, staging accuracy was reduced in patients with a history of prior intravesical and/or systemic chemotherapy. Conclusion: In staging bladder cancer, MRI, 11 C-acetate PET/CT and CT displayed similar levels of accuracy. For all modalities, a history of intravesical and/or systemic chemotherapy affected staging accuracy.

  4. Pretreatment evaluation of distant-site status in patients with nasopharyngeal carcinoma: accuracy of whole-body MRI at 3-Tesla and FDG-PET-CT

    International Nuclear Information System (INIS)

    Ng, Shu-Hang; Chan, Sheng-Chieh; Yen, Tzu-Chen; Chang, Joseph Tung-Chieh; Liao, Chun-Ta; Ko, Sheung-Fat; Wai, Yau-Yau; Wang, Hung-Ming; Wang, Jiun-Jie; Chen, Min-Chi

    2009-01-01

    We sought to prospectively evaluate the accuracy of 3.0-Tesla whole-body magnetic resonance imaging (WB-MRI) and integrated fluorodeoxyglucose (FDG) positron emission tomography (PET)/computed tomography (CT) (FDG-PET-CT), and their combined interpretation for the assessment of distant-site status in 150 patients with untreated nasopharyngeal carcinoma (NPC). Eighteen (12%) patients were diagnosed as having distant malignancies (15 patients had distant metastases, and three distant synchronous tumours). On a patient-based analysis, WB-MRI and FDG-PET-CT showed similar sensitivity (77.8% vs 72.2%, P > 0.999), specificity (98.5% vs 97.7%, P > 0.999) and diagnostic capability (0.905 vs 0.878, P = 0.669). Combined interpretation of WB-MRI and FDG-PET-CT showed no significant benefit over either technique alone. In conclusion, 3.0-Tesla WB-MRI is a feasible, non-ionising technique that showed similar diagnostic capacity to FDG-PET-CT in assessing distant-site status in patients with untreated NPC and can be recommended as the first-line imaging technique for comprehensive evaluation of such patients. (orig.)

  5. TU-AB-BRA-02: An Efficient Atlas-Based Synthetic CT Generation Method

    International Nuclear Information System (INIS)

    Han, X

    2016-01-01

    Purpose: A major obstacle for MR-only radiotherapy is the need to generate an accurate synthetic CT (sCT) from MR image(s) of a patient for the purposes of dose calculation and DRR generation. We propose here an accurate and efficient atlas-based sCT generation method, which has a computation speed largely independent of the number of atlases used. Methods: Atlas-based sCT generation requires a set of atlases with co-registered CT and MR images. Unlike existing methods that align each atlas to the new patient independently, we first create an average atlas and pre-align every atlas to the average atlas space. When a new patient arrives, we compute only one deformable image registration to align the patient MR image to the average atlas, which indirectly aligns the patient to all pre-aligned atlases. A patch-based non-local weighted fusion is performed in the average atlas space to generate the sCT for the patient, which is then warped back to the original patient space. We further adapt a PatchMatch algorithm that can quickly find top matches between patches of the patient image and all atlas images, which makes the patch fusion step also independent of the number of atlases used. Results: Nineteen brain tumour patients with both CT and T1-weighted MR images are used as testing data and a leave-one-out validation is performed. Each sCT generated is compared against the original CT image of the same patient on a voxel-by-voxel basis. The proposed method produces a mean absolute error (MAE) of 98.6±26.9 HU overall. The accuracy is comparable with a conventional implementation scheme, but the computation time is reduced from over an hour to four minutes. Conclusion: An average atlas space patch fusion approach can produce highly accurate sCT estimations very efficiently. Further validation on dose computation accuracy and using a larger patient cohort is warranted. The author is a full time employee of Elekta, Inc.

  6. Tree-structured vector quantization of CT chest scans: Image quality and diagnostic accuracy

    International Nuclear Information System (INIS)

    Cosman, P.C.; Tseng, C.; Gray, R.M.; Olshen, R.A.; Moses, L.E.; Davidson, H.C.; Bergin, C.J.; Riskin, E.A.

    1993-01-01

    The quality of lossy compressed images is often characterized by signal-to-noise ratios, informal tests of subjective quality, or receiver operating characteristic (ROC) curves that include subjective appraisals of the value of an image for a particular application. The authors believe that for medical applications, lossy compressed images should be judged by a more natural and fundamental aspect of relative image quality: their use in making accurate diagnoses. They apply a lossy compression algorithm to medical images, and quantify the quality of the images by the diagnostic performance of radiologists, as well as by traditional signal-to-noise ratios and subjective ratings. The study is unlike previous studies of the effects of lossy compression in that they consider non-binary detection tasks, simulate actual diagnostic practice instead of using paired tests or confidence rankings, use statistical methods that are more appropriate for non-binary clinical data than are the popular ROC curves, and use low-complexity predictive tree-structured vector quantization for compression rather than DCT-based transform codes combined with entropy coding. Their diagnostic tasks are the identification of nodules (tumors) in the lungs and lymphadenopathy in the mediastinum from computerized tomography (CT) chest scans. For the image modality, compression algorithm, and diagnostic tasks they consider, the original 12 bit per pixel (bpp) CT image can be compressed to between 1 bpp and 2 bpp with no significant changes in diagnostic accuracy

  7. The accuracy of {sup 18}F-FDG PET/CT in predicting the pathological response to neoadjuvant chemotherapy in patients with breast cancer. A meta-analysis and systematic review

    Energy Technology Data Exchange (ETDEWEB)

    Tian, Fangfang; Shen, Guohua; Diao, Wei; Jia, Zhiyun [West China Hospital of Sichuan University, Department of Nuclear Medicine, Chengdu, Sichuan (China); Deng, Yunfu [West China Hospital of Sichuan University, Department of Oncology, Chengdu (China)

    2017-11-15

    The aim of this meta-analysis was to evaluate the accuracy of {sup 18}F-FDG PET/CT in predicting the pathological response to neoadjuvant chemotherapy (NAC) in breast cancer (BC) patients. PubMed, Embase, the Cochrane Library (Central), and the Web of Science (SCI-Expanded) were systematically searched to identify pertinent studies. The methodologic quality of the included studies was assessed by the Quality Assessment of Diagnostic Accuracy Studies-2. The Spearman correlation coefficient was used to explore the existence of a threshold effect. Heterogeneity was assessed by the likelihood ratio I {sup 2} index. The pooled values calculated with a mixed-effects model for the sensitivity, specificity and diagnostic odds ratio with 95% confidence intervals were 81.9% (76.0-86.6%), 79.3% (72.1-85.1%) and 17.35 (10.98-27.42), respectively. {sup 18}F-FDG PET/CT has a moderate accuracy in predicting the pathological response during the early process of NAC in breast cancer patients. To increase the role of {sup 18}F-FDG PET/CT in monitoring the therapy response, future prospective studies are needed to explore how chemotherapy regimens and different subtypes affect the levels of glucose metabolism. (orig.)

  8. Low-dose non-enhanced CT versus full-dose contrast-enhanced CT in integrated PET/CT studies for the diagnosis of uterine cancer recurrence

    Energy Technology Data Exchange (ETDEWEB)

    Kitajima, Kazuhiro [Institute of Biomedical Research and Innovation, Department of PET Diagnosis, Kobe (Japan); Kobe University Graduate School of Medicine, Department of Radiology, Kobe (Japan); Suzuki, Kayo [Institute of Biomedical Research and Innovation, Department of PET Diagnosis, Kobe (Japan); Nakamoto, Yuji [Kyoto University Hospital, Department of Diagnostic Radiology, Kyoto (Japan); Onishi, Yumiko; Sakamoto, Setsu; Sugimura, Kazuro [Kobe University Graduate School of Medicine, Department of Radiology, Kobe (Japan); Senda, Michio [Institute of Biomedical Research and Innovation, Department of Molecular Imaging, Kobe (Japan); Kita, Masato [Kobe City Medical Center General Hospital, Department of Obstetrics and Gynecology, Kobe (Japan)

    2010-08-15

    To evaluate low-dose non-enhanced CT (ldCT) and full-dose contrast-enhanced CT (ceCT) in integrated {sup 18}F-fluorodeoxyglucose (FDG) PET/CT studies for restaging of uterine cancer. A group of 100 women who had undergone treatment for uterine cervical (n=55) or endometrial cancer (n=45) underwent a conventional PET/CT scans with ldCT, and then a ceCT scan. Two observers retrospectively reviewed and interpreted the PET/ldCT and PET/ceCT images in consensus using a three-point grading scale (negative, equivocal, or positive) per patient and per lesion. Final diagnoses were obtained by histopathological examination, or clinical follow-up for at least 6 months. Patient-based analysis showed that the sensitivity, specificity and accuracy of PET/ceCT were 90% (27/30), 97% (68/70) and 95% (95/100), respectively, whereas those of PET/ldCT were 83% (25/30), 94% (66/70) and 91% (91/100), respectively. Sensitivity, specificity and accuracy did not significantly differ between two methods (McNemar test, p=0.48, p=0.48, and p=0.13, respectively). There were 52 sites of lesion recurrence: 12 pelvic lymph node (LN), 11 local recurrence, 8 peritoneum, 7 abdominal LN, 5 lung, 3 supraclavicular LN, 3 liver, 2 mediastinal LN, and 1 muscle and bone. The grading results for the 52 sites of recurrence were: negative 5, equivocal 0 and positive 47 for PET/ceCT, and negative 5, equivocal 4 and positive 43 for PET/ldCT, respectively. Four equivocal regions by PET/ldCT (local recurrence, pelvic LN metastasis, liver metastasis and muscle metastasis) were correctly interpreted as positive by PET/ceCT. PET/ceCT is an accurate imaging modality for the assessment of uterine cancer recurrence. Its use reduces the frequency of equivocal interpretations. (orig.)

  9. SU-G-IeP2-15: Virtual Insertion of Digital Kidney Stones Into Dual-Source, Dual- Energy CT Projection Data

    International Nuclear Information System (INIS)

    Ferrero, A; Chen, B; Huang, A; Montoya, J; Yu, L; McCollough, C

    2016-01-01

    Purpose: In order to investigate novel methods to more accurately estimate the mineral composition of kidney stones using dual energy CT, it is desirable to be able to combine digital stones of known composition with actual phantom and patient scan data. In this work, we developed and validated a method to insert digital kidney stones into projection data acquired on a dual-source, dual-energy CT system. Methods: Attenuation properties of stones of different mineral composition were computed using tabulated mass attenuation coefficients, the chemical formula for each stone type, and the effective beam energy at each evaluated tube potential. A previously developed method to insert lesions into x-ray CT projection data was extended to include simultaneous dual-energy CT projections acquired on a dual-source gantry (Siemens Somatom Flash). Digital stones were forward projected onto both detectors and the resulting projections added to the physically acquired sinogram data. To validate the accuracy of the technique, digital stones were inserted into different locations in the ACR CT accreditation phantom; low and high contrast resolution, CT number accuracy and noise properties were compared before and after stone insertion. The procedure was repeated for two dual-energy tube potential pairs in clinical use on the scanner, 80/Sn140 kV and 100/Sn140 kV, respectively. Results: The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner. The largest average CT number difference for the 4 insert in the CT number accuracy module of the phantom was 3 HU. Conclusion: A framework was developed and validated for the creation of digital kidney stones of known mineral composition, and their projection-domain insertion into commercial dual-source, dual-energy CT projection data. This will allow a systematic investigation of the impact of scan and reconstruction parameters on stone attenuation and dual

  10. Performance evaluation of the CT component of the IRIS PET/CT preclinical tomograph

    Energy Technology Data Exchange (ETDEWEB)

    Panetta, Daniele [CNR Institute of Clinical Physiology (IFC-CNR), v. G. Moruzzi 1, I-56124 Pisa (Italy); Belcari, Nicola [Department of Physics “E. Fermi”, University of Pisa, L.go B. Pontecorvo 3, I-56127 Pisa (Italy); Tripodi, Maria [CNR Institute of Clinical Physiology (IFC-CNR), v. G. Moruzzi 1, I-56124 Pisa (Italy); Burchielli, Silvia [Fondazione CNR/Toscana “G. Monasterio” – FTGM, v. G. Moruzzi 1, I-56124 Pisa (Italy); Salvadori, Piero A. [CNR Institute of Clinical Physiology (IFC-CNR), v. G. Moruzzi 1, I-56124 Pisa (Italy); Del Guerra, Alberto [Department of Physics “E. Fermi”, University of Pisa, L.go B. Pontecorvo 3, I-56127 Pisa (Italy)

    2016-01-01

    In this paper, we evaluate the physical performance of the CT component of the IRIS scanner, a novel combined PET/CT scanner for preclinical imaging. The performance assessment is based on phantom measurement for the determination of image quality parameters (spatial resolution, linearity, geometric accuracy, contrast to noise ratio) and reproducibility in dynamic (4D) imaging. The CTDI{sub 100} has been measured free in air with a pencil ionization chamber, and the animal dose was calculated using Monte Carlo derived conversion factors taken from the literature. The spatial resolution at the highest quality protocol was 6.9 lp/mm at 10% of the MTF, using the smallest reconstruction voxel size of 58.8 μm. The accuracy of the reconstruction voxel size was within 0.1%. The linearity of the CT numbers as a function of the concentration of iodine was very good, with R{sup 2}>0.996 for all the tube voltages. The animal dose depended strongly on the scanning protocol, ranging from 158 mGy for the highest quality protocol (2 min, 80 kV) to about 12 mGy for the fastest protocol (7.3 s, 80 kV). In 4D dynamic modality, the maximum scanning rate reached was 3.1 frames per minute, using a short-scan protocol with 7.3 s of scan time per frame at the isotropic voxel size of 235 μm. The reproducibility of the system was high throughout the 10 frames acquired in dynamic modality, with a standard deviation of the CT values of all frames <8 HU and an average spatial reproducibility within 30% of the voxel size across all the field of view. Example images obtained during animal experiments are also shown.

  11. Performance evaluation of the CT component of the IRIS PET/CT preclinical tomograph

    Science.gov (United States)

    Panetta, Daniele; Belcari, Nicola; Tripodi, Maria; Burchielli, Silvia; Salvadori, Piero A.; Del Guerra, Alberto

    2016-01-01

    In this paper, we evaluate the physical performance of the CT component of the IRIS scanner, a novel combined PET/CT scanner for preclinical imaging. The performance assessment is based on phantom measurement for the determination of image quality parameters (spatial resolution, linearity, geometric accuracy, contrast to noise ratio) and reproducibility in dynamic (4D) imaging. The CTDI100 has been measured free in air with a pencil ionization chamber, and the animal dose was calculated using Monte Carlo derived conversion factors taken from the literature. The spatial resolution at the highest quality protocol was 6.9 lp/mm at 10% of the MTF, using the smallest reconstruction voxel size of 58.8 μm. The accuracy of the reconstruction voxel size was within 0.1%. The linearity of the CT numbers as a function of the concentration of iodine was very good, with R2>0.996 for all the tube voltages. The animal dose depended strongly on the scanning protocol, ranging from 158 mGy for the highest quality protocol (2 min, 80 kV) to about 12 mGy for the fastest protocol (7.3 s, 80 kV). In 4D dynamic modality, the maximum scanning rate reached was 3.1 frames per minute, using a short-scan protocol with 7.3 s of scan time per frame at the isotropic voxel size of 235 μm. The reproducibility of the system was high throughout the 10 frames acquired in dynamic modality, with a standard deviation of the CT values of all frames <8 HU and an average spatial reproducibility within 30% of the voxel size across all the field of view. Example images obtained during animal experiments are also shown.

  12. Performance evaluation of the CT component of the IRIS PET/CT preclinical tomograph

    International Nuclear Information System (INIS)

    Panetta, Daniele; Belcari, Nicola; Tripodi, Maria; Burchielli, Silvia; Salvadori, Piero A.; Del Guerra, Alberto

    2016-01-01

    In this paper, we evaluate the physical performance of the CT component of the IRIS scanner, a novel combined PET/CT scanner for preclinical imaging. The performance assessment is based on phantom measurement for the determination of image quality parameters (spatial resolution, linearity, geometric accuracy, contrast to noise ratio) and reproducibility in dynamic (4D) imaging. The CTDI_1_0_0 has been measured free in air with a pencil ionization chamber, and the animal dose was calculated using Monte Carlo derived conversion factors taken from the literature. The spatial resolution at the highest quality protocol was 6.9 lp/mm at 10% of the MTF, using the smallest reconstruction voxel size of 58.8 μm. The accuracy of the reconstruction voxel size was within 0.1%. The linearity of the CT numbers as a function of the concentration of iodine was very good, with R"2>0.996 for all the tube voltages. The animal dose depended strongly on the scanning protocol, ranging from 158 mGy for the highest quality protocol (2 min, 80 kV) to about 12 mGy for the fastest protocol (7.3 s, 80 kV). In 4D dynamic modality, the maximum scanning rate reached was 3.1 frames per minute, using a short-scan protocol with 7.3 s of scan time per frame at the isotropic voxel size of 235 μm. The reproducibility of the system was high throughout the 10 frames acquired in dynamic modality, with a standard deviation of the CT values of all frames <8 HU and an average spatial reproducibility within 30% of the voxel size across all the field of view. Example images obtained during animal experiments are also shown.

  13. SU-E-I-20: Comprehensive Quality Assurance Test of Second Generation Toshiba Aquilion Large Bore CT Simulator Based On AAPM TG-66 Recommendations

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, D [Toshiba America Medical Systems, Tustin, CA (United States)

    2015-06-15

    Purpose: AAPM radiation therapy committee task group No. 66 (TG-66) published a report which described a general approach to CT simulator QA. The report outlines the testing procedures and specifications for the evaluation of patient dose, radiation safety, electromechanical components, and image quality for a CT simulator. The purpose of this study is to thoroughly evaluate the performance of a second generation Toshiba Aquilion Large Bore CT simulator with 90 cm bore size (Toshiba, Nasu, JP) based on the TG-66 criteria. The testing procedures and results from this study provide baselines for a routine QA program. Methods: Different measurements and analysis were performed including CTDIvol measurements, alignment and orientation of gantry lasers, orientation of the tabletop with respect to the imaging plane, table movement and indexing accuracy, Scanogram location accuracy, high contrast spatial resolution, low contrast resolution, field uniformity, CT number accuracy, mA linearity and mA reproducibility using a number of different phantoms and measuring devices, such as CTDI phantom, ACR image quality phantom, TG-66 laser QA phantom, pencil ion chamber (Fluke Victoreen) and electrometer (RTI Solidose 400). Results: The CTDI measurements were within 20% of the console displayed values. The alignment and orientation for both gantry laser and tabletop, as well as the table movement and indexing and scanogram location accuracy were within 2mm as specified in TG66. The spatial resolution, low contrast resolution, field uniformity and CT number accuracy were all within ACR’s recommended limits. The mA linearity and reproducibility were both well below the TG66 threshold. Conclusion: The 90 cm bore size second generation Toshiba Aquilion Large Bore CT simulator that comes with 70 cm true FOV can consistently meet various clinical needs. The results demonstrated that this simulator complies with the TG-66 protocol in all aspects including electromechanical component

  14. Accuracy and reliability of different cone beam computed tomography (CBCT) devices for structural analysis of alveolar bone in comparison with multislice CT and micro-CT.

    Science.gov (United States)

    Van Dessel, Jeroen; Nicolielo, Laura Ferreira Pinheiro; Huang, Yan; Coudyzer, Walter; Salmon, Benjamin; Lambrichts, Ivo; Jacobs, Reinhilde

    for the i-CAT Next Generation CBCT (ICC 0.26 min to 0.86 max) and standard resolution MSCT (ICC 0.10 min to 0.62 max). This study demonstrated that most CBCT machines may be able to quantitatively assess alveolar bone quality, with a level of accuracy and reliability that approaches micro-CT. One may therefore propose to extrapolate this to clinical CBCT imaging, certainly when there is a need for implant rehabilitation in dentate jaw bones. Conflict-of-interest statement: There is no conflict of interest to declare. Fellowship support was received from Research Foundation Flanders (FWO) from the Belgian government and from the Coordination for the Improvement of Higher Education Personnel (CAPES) programme, Science without Borders, from the Brazilian government.

  15. SU-F-T-403: Impact of Dose Reduction for Simulation CT On Radiation Therapy Treatment Planning

    Energy Technology Data Exchange (ETDEWEB)

    Liang, Q; Shah, P; Li, S; Miyamoto, C [Temple University Hospital, Philadelphia, PA (United States)

    2016-06-15

    Purpose: To investigate the feasibility of applying ALARA principles to current treatment planning CT scans. The study aims to quantitatively verify lower dose scans does not alter treatment planning. Method: Gammex 467 tissue characterization phantom with inserts of 14 different materials was scanned at seven different mA levels (30∼300 mA). CT numbers of different inserts were measured. Auto contouring for bone and lung in treatment planning system (Pinnacle) was used to evaluate the effect of CT number accuracy from treatment planning aspect, on the 30 and 300 mA-scanned images. A head CT scan intended for a 3D whole brain radiation treatment was evaluated. Dose calculations were performed on normal scanned images using clinical protocol (120 kVP, Smart mA, maximum 291 mA), and the images with added simulating noise mimicking a 70 mA scan. Plan parameters including isocenter, beam arrangements, block shapes, dose grid size and resolution, and prescriptions were kept the same for these two plans. The calculated monitor units (MUs) for these two plans were compared. Results: No significant degradation of CT number accuracy was found at lower dose levels from both the phantom scans, and the patient images with added noise. The CT numbers kept consistent when mA is higher than 60 mA. The auto contoured volumes for lung and cortical bone show 0.3% and 0.12% of differences between 30 mA and 300 mA respectively. The two forward plans created on regular and low dose images gave the same calculated MU, and 98.3% of points having <1% of dose difference. Conclusion: Both phantom and patient studies quantitatively verified low dose CT provides similar quality for treatment planning at 20–25% of regular scan dose. Therefore, there is the potential to optimize simulation CT scan protocol to fulfil the ALARA principle and limit unnecessary radiation exposure to non-targeted tissues.

  16. MR and CT for malignant hypernephromas

    International Nuclear Information System (INIS)

    Uhlenbrock, D.; Fischer, C.; Ruehl, G.; Beyer, H.K.; Hummelsheim, P.; Bochum Univ.; Marienhospital, Herne; Bochum Univ.

    1987-01-01

    Forty patients with histologically confirmed hypernephromas were examined by MR and CT. As regards T-staging, MR had an accuracy of 97 % and CT of 91%; for N-staging, MR had an accuracy of 85% and CT of 91% and, with regard to infiltration of the renal vein, MR was correct in 88% of cases and CT in 81%. MR was most accurate in the pre-operative assessment of tumour spread, demonstrating caval involvement in 100% and lymph node metastases in 97%. On the other hand, infiltration of the renal pelvis could be assessed in only 65% of cases; in 27% the assessment was not possible and in 8% one could not be certain. CT proved to be at its best for staging of lymph nodes, but in 15% it was not possible to identify infiltration of the renal vein. MR was not greatly superior to CT in the preoperative diagnosis and staging of hypernephromas. (orig.) [de

  17. Prospective evaluation of MRI, {sup 11}C-acetate PET/CT and contrast-enhanced CT for staging of bladder cancer

    Energy Technology Data Exchange (ETDEWEB)

    Vargas, H.A., E-mail: vargasah@mskcc.org [Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); Akin, O.; Schöder, H. [Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); Olgac, S. [Department of Pathology, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); Dalbagni, G. [Department of Surgery, Urology Service, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); Hricak, H. [Department of Radiology, Memorial Sloan-Kettering Cancer Center, New York, NY (United States); Bochner, B.H. [Department of Surgery, Urology Service, Memorial Sloan-Kettering Cancer Center, New York, NY (United States)

    2012-12-15

    Purpose: To prospectively evaluate the diagnostic performance of magnetic resonance imaging (MRI), {sup 11}C-acetate positron emission tomography/computed tomography (PET/CT) and contrast-enhanced CT for bladder cancer staging, using whole-mount pathologic review of radical cystectomy and pelvic lymph node specimens as the reference standard. Materials and methods: The institutional review board approved this prospective study, which was compliant with the Health Insurance Portability and Accountability Act. Written informed consent was obtained from 16 patients with histologically confirmed bladder cancer who underwent MRI, {sup 11}C-acetate PET/CT and contrast-enhanced CT before radical cystectomy and pelvic lymph node dissection. Before imaging 4/16 patients had received intravesical Bacillus Calmette-Guérin treatment, 6 had received systemic chemotherapy, 3 had received both and 3 had received neither. Measures of diagnostic performance including accuracy, sensitivity and specificity were estimated separately for each imaging modality. Results: MRI correctly staged 56% of patients (9/16), overstaged 38% (6/16) and understaged 6% (1/16). CT correctly staged 50% of patients (8/16), overstaged 44% (7/16) and understaged 6% (1/16). In 9 patients, {sup 11}C-acetate PET/CT showed uptake within the bladder wall; the uptake was true-positive in 7 patients and false-positive in 2 patients. Of the remaining 7 patients, 5 had true-negative and 2 had false-negative PET/CT results for cancer in the bladder wall. For all modalities, staging accuracy was reduced in patients with a history of prior intravesical and/or systemic chemotherapy. Conclusion: In staging bladder cancer, MRI, {sup 11}C-acetate PET/CT and CT displayed similar levels of accuracy. For all modalities, a history of intravesical and/or systemic chemotherapy affected staging accuracy.

  18. Accuracy of Estimation of Graft Size for Living-Related Liver Transplantation: First Results of a Semi-Automated Interactive Software for CT-Volumetry

    Science.gov (United States)

    Mokry, Theresa; Bellemann, Nadine; Müller, Dirk; Lorenzo Bermejo, Justo; Klauß, Miriam; Stampfl, Ulrike; Radeleff, Boris; Schemmer, Peter; Kauczor, Hans-Ulrich; Sommer, Christof-Matthias

    2014-01-01

    Objectives To evaluate accuracy of estimated graft size for living-related liver transplantation using a semi-automated interactive software for CT-volumetry. Materials and Methods Sixteen donors for living-related liver transplantation (11 male; mean age: 38.2±9.6 years) underwent contrast-enhanced CT prior to graft removal. CT-volumetry was performed using a semi-automated interactive software (P), and compared with a manual commercial software (TR). For P, liver volumes were provided either with or without vessels. For TR, liver volumes were provided always with vessels. Intraoperative weight served as reference standard. Major study goals included analyses of volumes using absolute numbers, linear regression analyses and inter-observer agreements. Minor study goals included the description of the software workflow: degree of manual correction, speed for completion, and overall intuitiveness using five-point Likert scales: 1–markedly lower/faster/higher for P compared with TR, 2–slightly lower/faster/higher for P compared with TR, 3–identical for P and TR, 4–slightly lower/faster/higher for TR compared with P, and 5–markedly lower/faster/higher for TR compared with P. Results Liver segments II/III, II–IV and V–VIII served in 6, 3, and 7 donors as transplanted liver segments. Volumes were 642.9±368.8 ml for TR with vessels, 623.8±349.1 ml for P with vessels, and 605.2±345.8 ml for P without vessels (Pvolumetry performed with P can predict accurately graft size for living-related liver transplantation while improving workflow compared with TR. PMID:25330198

  19. Efficacy of IV Buscopan as a muscle relaxant in CT colonography

    International Nuclear Information System (INIS)

    Bruzzi, John F.; Brennan, Darren D.; Fenlon, Helen M.; Moss, Alan C.; MacMathuna, Padraic

    2003-01-01

    The aim of this study was to examine the efficacy of IV Buscopan as a muscle relaxant in CT colonography in terms of colonic distension and polyp detection, and to determine its particular efficacy in patients with diverticular disease. Seventy-three consecutive patients were randomised to receive IV Buscopan or no muscle relaxant prior to CT colonography. CT colonography was performed using a Siemens Somatom 4-detector multislice CT scanner. The following parameters were recorded: degree of colonic distension using a 4-point scale; diagnostic adequacy of colonic distension; presence or absence of diverticular disease; and presence of colonic polyps. Accuracy of polyp detection was assessed using subsequent conventional colonoscopy as a gold standard. There was no significant difference between the two groups in the number of segments that were deemed to be optimally or adequately distended (p=0.37). Although IV Buscopan did improve distension of certain segments, this effect was not sufficient to improve the number of diagnostically adequate studies in the Buscopan group (p=0.14). In patients with diverticular disease, IV Buscopan did not have any significant effect on segments affected by diverticulosis but was associated with an improvement in distension of more proximal segments. There was no significant difference between the two groups in terms of polyp detection (p=0.34). The addition of prone scanning to supine scanning was found to be the most useful technique for maximising colonic distension. Intravenous Buscopan at CT colonography does not improve the overall adequacy of colonic distension nor the accuracy of polyp detection. In patients with sigmoid diverticular disease IV Buscopan improves distension of more proximal colonic segments and may be useful in selected cases, but our results do not support its routine use for CT colonography. (orig.)

  20. Diagnostic yield and accuracy of CT angiography, MR angiography, and digital subtraction angiography for detection of macrovascular causes of intracerebral haemorrhage: Prospective, multicentre cohort study

    NARCIS (Netherlands)

    C.J.J. Van Asch (Charlotte J.J.); B.K. Velthuis (Birgitta K.); G.J.E. Rinkel (Gabriël J.E.); A. Algra (Ale); G.A.P. de Kort (G. A P); T.D. Witkamp (Theo); J.C.M. De Ridder (Johanna C.M.); K.M. Van Nieuwenhuizen (Koen M.); F.-E. De Leeuw (Frank-Erik); W.J. Schonewille (Wouter); P.L.M. de Kort (Paul); D.W.J. Dippel (Diederik); T.W.M. Raaymakers (Theodora W.M.); J. Hofmeijer; M.J.H. Wermer (Marieke); H. Kerkhoff (Henk); K. Jellema (Korné); I.M. Bronner (Irene M.); M.J.M. Remmers (Michel ); H.P. Bienfait (Henri); R.J.G.M. Witjes (Ron J.G.M.); J.P. Greving (Jacoba); C.J.M. Klijn (Catharina J.M.); H.F. de Leeuw (Frank); H.B. Boogaarts; E.J. van Dijk (Ewoud); W.J. Schonewille; W.M.J. Pellikaan; C. Puppels-De Waard; P.L.M. De Kort; J.P. Peluso; J. van Tuijl (Jordie); J. Hofmeijer; F.B.M. Joosten (Frank); D.W.J. Dippel (Diederik); L. Khajeh (Ladbon); T.W.M. Raaijmakers; M.J. Wermer; M.A.A. van Walderveen (Marianne); H. Kerkhoff; E. Zock; K. Jellema (Korné); G.J. Lycklama à Nijeholt (Geert); I.M. Bronner; M.J.M. Remmers; R.J.G.M. Witjes; H.P. Bienfait; K.E. Droogh-Greve; R. Donders (Rogier); V.I.H. Kwa; T.H.C.M.L. Schreuder (Tobien H. C. M. L.); C.L. Franke (Cees); J.S. Straver; C. Jansen; S.L.M. Bakker (Stef); C.C. Pleiter (C.); M.C. Visser; C.J.J. Van Asch; B.K. Velthuis (Birgitta); G.J.E. Rinkel (Gabriel); K.M. Van Nieuwenhuizen; C.J.M. Klijn (Catharina J.M.)

    2015-01-01

    textabstractStudy question What are the diagnostic yield and accuracy of early computed tomography (CT) angiography followed by magnetic resonance imaging/angiography (MRI/MRA) and digital subtraction angiography (DSA) in patients with non-traumatic intracerebral haemorrhage? Methods This

  1. Clinical evaluation of a commercial orthopedic metal artifact reduction tool for CT simulations in radiation therapy

    Energy Technology Data Exchange (ETDEWEB)

    Li Hua; Noel, Camille; Chen, Haijian; Harold Li, H.; Low, Daniel; Moore, Kevin; Klahr, Paul; Michalski, Jeff; Gay, Hiram A.; Thorstad, Wade; Mutic, Sasa [Department of Radiation Oncology, Washington University, St. Louis, Missouri 63110 (United States); Department of Radiation Oncology, University of California Los Angeles, Los Angeles, California 90095 (United States); Department of Radiation Oncology, University of California San Diego, San Diego, California 92093 (United States); Philips Healthcare System, Cleveland, Ohio 44143 (United States); Department of Radiation Oncology, Washington University, St. Louis, Missouri 63110 (United States)

    2012-12-15

    Purpose: Severe artifacts in kilovoltage-CT simulation images caused by large metallic implants can significantly degrade the conspicuity and apparent CT Hounsfield number of targets and anatomic structures, jeopardize the confidence of anatomical segmentation, and introduce inaccuracies into the radiation therapy treatment planning process. This study evaluated the performance of the first commercial orthopedic metal artifact reduction function (O-MAR) for radiation therapy, and investigated its clinical applications in treatment planning. Methods: Both phantom and clinical data were used for the evaluation. The CIRS electron density phantom with known physical (and electron) density plugs and removable titanium implants was scanned on a Philips Brilliance Big Bore 16-slice CT simulator. The CT Hounsfield numbers of density plugs on both uncorrected and O-MAR corrected images were compared. Treatment planning accuracy was evaluated by comparing simulated dose distributions computed using the true density images, uncorrected images, and O-MAR corrected images. Ten CT image sets of patients with large hip implants were processed with the O-MAR function and evaluated by two radiation oncologists using a five-point score for overall image quality, anatomical conspicuity, and CT Hounsfield number accuracy. By utilizing the same structure contours delineated from the O-MAR corrected images, clinical IMRT treatment plans for five patients were computed on the uncorrected and O-MAR corrected images, respectively, and compared. Results: Results of the phantom study indicated that CT Hounsfield number accuracy and noise were improved on the O-MAR corrected images, especially for images with bilateral metal implants. The {gamma} pass rates of the simulated dose distributions computed on the uncorrected and O-MAR corrected images referenced to those of the true densities were higher than 99.9% (even when using 1% and 3 mm distance-to-agreement criterion), suggesting that dose

  2. Hybrid Brain–Computer Interface Techniques for Improved Classification Accuracy and Increased Number of Commands: A Review

    Science.gov (United States)

    Hong, Keum-Shik; Khan, Muhammad Jawad

    2017-01-01

    In this article, non-invasive hybrid brain–computer interface (hBCI) technologies for improving classification accuracy and increasing the number of commands are reviewed. Hybridization combining more than two modalities is a new trend in brain imaging and prosthesis control. Electroencephalography (EEG), due to its easy use and fast temporal resolution, is most widely utilized in combination with other brain/non-brain signal acquisition modalities, for instance, functional near infrared spectroscopy (fNIRS), electromyography (EMG), electrooculography (EOG), and eye tracker. Three main purposes of hybridization are to increase the number of control commands, improve classification accuracy and reduce the signal detection time. Currently, such combinations of EEG + fNIRS and EEG + EOG are most commonly employed. Four principal components (i.e., hardware, paradigm, classifiers, and features) relevant to accuracy improvement are discussed. In the case of brain signals, motor imagination/movement tasks are combined with cognitive tasks to increase active brain–computer interface (BCI) accuracy. Active and reactive tasks sometimes are combined: motor imagination with steady-state evoked visual potentials (SSVEP) and motor imagination with P300. In the case of reactive tasks, SSVEP is most widely combined with P300 to increase the number of commands. Passive BCIs, however, are rare. After discussing the hardware and strategies involved in the development of hBCI, the second part examines the approaches used to increase the number of control commands and to enhance classification accuracy. The future prospects and the extension of hBCI in real-time applications for daily life scenarios are provided. PMID:28790910

  3. Hybrid Brain-Computer Interface Techniques for Improved Classification Accuracy and Increased Number of Commands: A Review.

    Science.gov (United States)

    Hong, Keum-Shik; Khan, Muhammad Jawad

    2017-01-01

    In this article, non-invasive hybrid brain-computer interface (hBCI) technologies for improving classification accuracy and increasing the number of commands are reviewed. Hybridization combining more than two modalities is a new trend in brain imaging and prosthesis control. Electroencephalography (EEG), due to its easy use and fast temporal resolution, is most widely utilized in combination with other brain/non-brain signal acquisition modalities, for instance, functional near infrared spectroscopy (fNIRS), electromyography (EMG), electrooculography (EOG), and eye tracker. Three main purposes of hybridization are to increase the number of control commands, improve classification accuracy and reduce the signal detection time. Currently, such combinations of EEG + fNIRS and EEG + EOG are most commonly employed. Four principal components (i.e., hardware, paradigm, classifiers, and features) relevant to accuracy improvement are discussed. In the case of brain signals, motor imagination/movement tasks are combined with cognitive tasks to increase active brain-computer interface (BCI) accuracy. Active and reactive tasks sometimes are combined: motor imagination with steady-state evoked visual potentials (SSVEP) and motor imagination with P300. In the case of reactive tasks, SSVEP is most widely combined with P300 to increase the number of commands. Passive BCIs, however, are rare. After discussing the hardware and strategies involved in the development of hBCI, the second part examines the approaches used to increase the number of control commands and to enhance classification accuracy. The future prospects and the extension of hBCI in real-time applications for daily life scenarios are provided.

  4. Hybrid Brain–Computer Interface Techniques for Improved Classification Accuracy and Increased Number of Commands: A Review

    Directory of Open Access Journals (Sweden)

    Keum-Shik Hong

    2017-07-01

    Full Text Available In this article, non-invasive hybrid brain–computer interface (hBCI technologies for improving classification accuracy and increasing the number of commands are reviewed. Hybridization combining more than two modalities is a new trend in brain imaging and prosthesis control. Electroencephalography (EEG, due to its easy use and fast temporal resolution, is most widely utilized in combination with other brain/non-brain signal acquisition modalities, for instance, functional near infrared spectroscopy (fNIRS, electromyography (EMG, electrooculography (EOG, and eye tracker. Three main purposes of hybridization are to increase the number of control commands, improve classification accuracy and reduce the signal detection time. Currently, such combinations of EEG + fNIRS and EEG + EOG are most commonly employed. Four principal components (i.e., hardware, paradigm, classifiers, and features relevant to accuracy improvement are discussed. In the case of brain signals, motor imagination/movement tasks are combined with cognitive tasks to increase active brain–computer interface (BCI accuracy. Active and reactive tasks sometimes are combined: motor imagination with steady-state evoked visual potentials (SSVEP and motor imagination with P300. In the case of reactive tasks, SSVEP is most widely combined with P300 to increase the number of commands. Passive BCIs, however, are rare. After discussing the hardware and strategies involved in the development of hBCI, the second part examines the approaches used to increase the number of control commands and to enhance classification accuracy. The future prospects and the extension of hBCI in real-time applications for daily life scenarios are provided.

  5. FDG-PET/CT in the diagnosis of recurrent breast cancer

    International Nuclear Information System (INIS)

    Murakami, Ryusuke; Kumita, Shin-ichiro; Yoshida, Tamiko; Ishihara, Keiichi; Kiriyama, Tomonari; Hakozaki, Kenta; Yanagihara, Keiko; Lida, Shinya; Tsuchiya, Shin-ichi

    2012-01-01

    Background. An advantage of PET/CT has been demonstrated for diagnosis of several tumor entities. In patients with breast cancer, early diagnosis and accurate restaging of recurrence after surgery is important for selection of the most appropriate therapeutic strategy. Purpose. To evaluate the accuracy of integrated positron emission tomography and computed tomography (PET/CT) using 18F-fluorodeoxyglucose (FDG), for follow-up of patients with suspected recurrent breast cancer. Material and Methods. Forty-seven patients with suspected recurrent breast cancer underwent PET/CT. The PET and PET/CT images were interpreted without knowledge of the results of other diagnostic modalities, and compared with each other with reference to the final diagnosis. Results. Twenty-five (53%) patients suffered tumor recurrence. The overall sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of PET/CT were 96%, 91%, 92%, 95%, and 94%, respectively. In comparison with PET, PET/CT had a higher sensitivity and accuracy (96% vs. 80% and 94% vs. 81%, respectively). The difference in diagnostic accuracy between PET/CT and PET was significant (P < 0.05). Conclusion. The present findings indicate that PET/CT is an accurate, sensitive and reliable modality for screening and detection of breast cancer recurrence. PET/CT appears to be an effective surveillance tool, as it is able to cover the whole body in a single procedure and shows good performance

  6. 18F-Fluorocholine PET/CT as a complementary tool in the follow-up of low-grade glioma: diagnostic accuracy and clinical utility

    International Nuclear Information System (INIS)

    Gomez-Rio, Manuel; Rodriguez-Fernandez, Antonio; Llamas-Elvira, Jose M.; Testart Dardel, Nathalie; Santiago Chinchilla, Alicia; Olivares Granados, Gonzalo; Luque Caro, Raquel; Zurita Herrera, Mercedes; Chamorro Santos, Clara E.; Lardelli-Claret, Pablo

    2015-01-01

    The follow-up of treated low-grade glioma (LGG) requires the evaluation of subtle clinical changes and MRI results. When the result is inconclusive, additional procedures are required to assist decision-making, such as the use of advanced MRI (aMRI) sequences and nuclear medicine scans (SPECT and PET). The aim of this study was to determine whether incorporating 18 F-fluorocholine PET/CT in the follow-up protocol for treated LGG improves diagnostic accuracy and clinical utility. This was a prospective case-series study in patients with treated LGG during standard follow-up with indeterminate clinical and/or radiological findings of tumour activity. All patients underwent clinical evaluation, aMRI, 201 Tl-SPECT and 18 F-fluorocholine PET/CT. Images were interpreted by visual evaluation complemented with semiquantitative analysis. Between January 2012 and December 2013, 18 patients were included in this study. The final diagnosis was established by histology (five surgical specimens, one biopsy specimen) or by consensus of the Neuro-Oncology Group (11 patients) after a follow-up of >6 months (mean 14.9 ± 2.72 months). The global diagnostic accuracies were 90.9 % for aMRI (38.8 % inconclusive), 69.2 % for 201 Tl-SPECT (11.1 % inconclusive), and 100 % for 18 F-fluorocholine PET/CT. 201 Tl-SPECT led correctly to a change in the initial approach in 38.9 % of patients but might have led to error in 27.8 %. The use of 18 F-fluorocholine PET/CT alone rather than 201 Tl-SPECT led correctly to a change in the approach suggested by routine follow-up in 72.2 % of patients and endorsed the approach in the remaining 27.8 %. Our results support the need to complement structural MRI with aMRI and nuclear medicine procedures in selected patients. 18 F-Fluorocholine PET/CT can be useful in the individualized management of patients with treated LGG with uncertain clinical and/or radiological evidence of tumour activity. (orig.)

  7. SU-E-J-272: Auto-Segmentation of Regions with Differentiating CT Numbers for Treatment Response Assessment

    International Nuclear Information System (INIS)

    Yang, C; Noid, G; Dalah, E; Paulson, E; Li, X; Gilat-Schmidt, T

    2015-01-01

    Purpose: It has been reported recently that the change of CT number (CTN) during and after radiation therapy (RT) may be used to assess RT response. The purpose of this work is to develop a tool to automatically segment the regions with differentiating CTN and/or with change of CTN in a series of CTs. Methods: A software tool was developed to identify regions with differentiating CTN using K-mean Cluster of CT numbers and to automatically delineate these regions using convex hull enclosing method. Pre- and Post-RT CT, PET, or MRI images acquired for sample lung and pancreatic cancer cases were used to test the software tool. K-mean cluster of CT numbers within the gross tumor volumes (GTVs) delineated based on PET SUV (standard uptake value of fludeoxyglucose) and/or MRI ADC (apparent diffusion coefficient) map was analyzed. The cluster centers with higher value were considered as active tumor volumes (ATV). The convex hull contours enclosing preset clusters were used to delineate these ATVs with color washed displays. The CTN defined ATVs were compared with the SUV- or ADC-defined ATVs. Results: CTN stability of the CT scanner used to acquire the CTs in this work is less than 1.5 Hounsfield Unit (HU) variation annually. K-mean cluster centers in the GTV have difference of ∼20 HU, much larger than variation due to CTN stability, for the lung cancer cases studied. The dice coefficient between the ATVs delineated based on convex hull enclosure of high CTN centers and the PET defined GTVs based on SUV cutoff value of 2.5 was 90(±5)%. Conclusion: A software tool was developed using K-mean cluster and convex hull contour to automatically segment high CTN regions which may not be identifiable using a simple threshold method. These CTN regions were reasonably overlapped with the PET or MRI defined GTVs

  8. CT vs. MRCP in choledocholithiasis jaundice.

    Science.gov (United States)

    Petrescu, I; Bratu, A M; Petrescu, S; Popa, B V; Cristian, D; Burcos, T

    2015-01-01

    Obstructive jaundice can raise problems to diagnostic imaging. The radiologist must choose the most appropriate examination that delivers the most important diagnostic information because the differences between a lithiasic obstruction and a tumoral one are vital. This information helps the surgeon speed up the process of decision-making, because the treatment may be very different in relation to the nature of the obstruction. This study tries to demonstrate the diagnostic accuracy of computed tomography (CT) and magnetic resonance cholangiopancreatography (MRCP) in detecting the obstacle in the common bile duct (CBD) and the possibility of establishing the lithiasic nature of the obstruction. A retrospective analysis was analyzed during an interval of 18 months that included jaundice patients admitted in the General Surgery Department of "Coltea" Clinical Hospital. They were examined by CT scanning and by MRCP, being suspected of choledocholithiasis. 63 patients were included in the study, 34 females and 29 males. 33 CT scans and 30 MRCP exams were performed. CT scan is useful in detecting residual or iterative choledocholithiasis in patients after cholecystectomy, contrast enhanced CT (CECT), being able to differentiate between lithiasic and non-lithiasic obstruction. MRCP delivers important anatomic details of the biliary tree; it is superior to CT in diagnosing the hepatocholedochal lithiasis; MRCP tends to replace endoscopic retrograde cholangiopancreatography (ERCP)--the diagnostic "gold standard" reducing the number of unnecessary invasive diagnostic procedures.

  9. SU-C-18A-05: Registration Accuracy of MR-Based Images to On-Board Megavoltage Cone-Beam CT for Brain Patient Setup

    International Nuclear Information System (INIS)

    Pinnaduwage, D S; Chen, J; Descovich, M; Pouliot, J; Hwang, Ken-Ping

    2014-01-01

    Purpose: To quantify the difference in isocenter shifts when co-registering MR and MR-based pseudo CTs (pCT) with on-board megavoltage conebeam CT (CBCT) images. Methods: Fast Spoiled Gradient Echo MRs were used to generate pCTs (research version of Advantage Sim MD™, GE Healthcare) for ten patients who had prior brain radiotherapy. The planning CT (rCT) for each was co-registered with the MR, and the plan isocenter and two other reference points were transferred to the MR and pCT. CBCT images (with the machine isocenter) from a single treatment day were coregistered with the 3 test images (MR, pCT and rCT), by two observers and by an automated registration algorithm. The reference points were used to calculate patient shifts and rotations from the registrations. The shifts calculated from the test image registrations were compared to each other and to the shifts performed by the therapists who treated the patients on that day. Results: The average difference in absolute value between the isocenter shifts from the MR-, pCT- and rCT-CBCT registrations, and the therapist shifts, were 2.02, 3.01 and 0.89 mm (craniocaudal), 1.14, 1.34 and 0.46 mm (lateral), and 1.37, 3.43 and 1.43 mm (vertical), respectively. The MR- and pCT-CBCT registrations differed by 1.99, and 2.53 mm (craniocaudal), 1.36, and 1.37 mm (lateral), and 0.74 and 2.34 mm (vertical), respectively, from the average rCT-CBCT shifts. On average, differences of 2.39 (craniocaudal), 1.28 (lateral) and 2.84 mm (vertical) were seen between the MR and pCT shifts. Rotations relative to the CBCT coordinate system were on average <2° for the MR and rCT, and <6° for the pCT. Conclusion: In this study, FSPGR MR-CBCT registrations were more precise compared to the pCT-CBCT registrations. For improved accuracy, MR sequences that are optimal for bony anatomy visualization are necessary. GE healthcare has provided a research version of Advantage Sim MD to UCSF. No financial support was provided

  10. Material elemental decomposition in dual and multi-energy CT via a sparsity-dictionary approach for proton stopping power ratio calculation.

    Science.gov (United States)

    Shen, Chenyang; Li, Bin; Chen, Liyuan; Yang, Ming; Lou, Yifei; Jia, Xun

    2018-04-01

    Accurate calculation of proton stopping power ratio (SPR) relative to water is crucial to proton therapy treatment planning, since SPR affects prediction of beam range. Current standard practice derives SPR using a single CT scan. Recent studies showed that dual-energy CT (DECT) offers advantages to accurately determine SPR. One method to further improve accuracy is to incorporate prior knowledge on human tissue composition through a dictionary approach. In addition, it is also suggested that using CT images with multiple (more than two) energy channels, i.e., multi-energy CT (MECT), can further improve accuracy. In this paper, we proposed a sparse dictionary-based method to convert CT numbers of DECT or MECT to elemental composition (EC) and relative electron density (rED) for SPR computation. A dictionary was constructed to include materials generated based on human tissues of known compositions. For a voxel with CT numbers of different energy channels, its EC and rED are determined subject to a constraint that the resulting EC is a linear non-negative combination of only a few tissues in the dictionary. We formulated this as a non-convex optimization problem. A novel algorithm was designed to solve the problem. The proposed method has a unified structure to handle both DECT and MECT with different number of channels. We tested our method in both simulation and experimental studies. Average errors of SPR in experimental studies were 0.70% in DECT, 0.53% in MECT with three energy channels, and 0.45% in MECT with four channels. We also studied the impact of parameter values and established appropriate parameter values for our method. The proposed method can accurately calculate SPR using DECT and MECT. The results suggest that using more energy channels may improve the SPR estimation accuracy. © 2018 American Association of Physicists in Medicine.

  11. Pancreatic neuroendocrine tumour (PNET): Staging accuracy of MDCT and its diagnostic performance for the differentiation of PNET with uncommon CT findings from pancreatic adenocarcinoma

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Jung Hoon; Lee, Jeong Min; Han, Joon Koo; Choi, Byung-Ihn [Seoul National University Hospital, Department of Radiology, 101 Daehangno, Jongno-gu, Seoul (Korea, Republic of); Seoul National University College of Medicine, Institute of Radiation Medicine, Seoul (Korea, Republic of); Eun, Hyo Won [Yonsei University College of Medicine, Department of Radiology, Severance Hospital, Seodaemun-ku, Seoul (Korea, Republic of); Kim, Young Jae [Soonchunhyang University Hospital, Department of Radiology, 657 Hannam-Dong, Youngsan-Ku, Seoul (Korea, Republic of)

    2016-05-15

    To investigate staging accuracy of multidetector CT (MDCT) for pancreatic neuroendocrine tumour (PNET) and diagnostic performance for differentiation of PNET from pancreatic adenocarcinoma. We included 109 patients with surgically proven PNET (NETG1 = 66, NETG2 = 31, NEC = 12) who underwent MDCT. Two reviewers assessed stage and presence of predefined CT findings. We analysed the relationship between CT findings and tumour grade. Using PNETs with uncommon findings, we also estimated the possibility of PNET or adenocarcinoma. Accuracy for T stage was 85-88 % and N-metastasis was 83-89 %. Common findings included well circumscribed, homogeneously enhanced, hypervascular mass, common in lower grade tumours (p < 0.05). Uncommon findings included ill-defined, heterogeneously enhanced, hypovascular mass and duct dilation, common in higher grade tumours (p < 0.05). Using 31 PNETs with uncommon findings, diagnostic performance for differentiation from adenocarcinoma was 0.760-0.806. Duct dilatation was an independent predictor for adenocarcinoma (Exp(B) = 4.569). PNETs with uncommon findings were associated with significantly worse survival versus PNET with common findings (62.7 vs. 95.7 months, p < 0.001). MDCT is useful for preoperative evaluation of PNET; it not only accurately depicts the tumour stage but also prediction of tumour grade, because uncommon findings were more common in higher grade tumours. (orig.)

  12. Value and Accuracy of Multidetector Computed Tomography in Obstructive Jaundice

    International Nuclear Information System (INIS)

    Mathew, Rishi Philip; Moorkath, Abdunnisar; Basti, Ram Shenoy; Suresh, Hadihally B.

    2016-01-01

    Objective; To find out the role of MDCT in the evaluation of obstructive jaundice with respect to the cause and level of the obstruction, and its accuracy. To identify the advantages of MDCT with respect to other imaging modalities. To correlate MDCT findings with histopathology/surgical findings/Endoscopic Retrograde CholangioPancreatography (ERCP) findings as applicable. This was a prospective study conducted over a period of one year from August 2014 to August 2015. Data were collected from 50 patients with clinically suspected obstructive jaundice. CT findings were correlated with histopathology/surgical findings/ERCP findings as applicable. Among the 50 people studied, males and females were equal in number, and the majority belonged to the 41–60 year age group. The major cause for obstructive jaundice was choledocholithiasis. MDCT with reformatting techniques was very accurate in picking a mass as the cause for biliary obstruction and was able to differentiate a benign mass from a malignant one with high accuracy. There was 100% correlation between the CT diagnosis and the final diagnosis regarding the level and type of obstruction. MDCT was able to determine the cause of obstruction with an accuracy of 96%. MDCT with good reformatting techniques has excellent accuracy in the evaluation of obstructive jaundice with regards to the level and cause of obstruction

  13. Evolution of overall CT in the staging of patients with Hodgkin's disease

    International Nuclear Information System (INIS)

    Maurizi Enrici, R.; Capua, A.; Osti, M.F.; Sarra, R.; Tombolini, V.; Anselmo, A.P.; Campodonico, F.

    1991-01-01

    From 1983 through 1989, 141 untreated patients with Hodgkin's disease underwent CT of the abdomen. They subsequently underwent staging laparotomy plus splenectomy and multiple biopsies of liver and lymph nodes, at the Institute of Radiology and Hematology, University 'La Sapienza', Rome. CT results were compared with surgical findings to evaluate CT sensitivity, specificity, and overall accuracy. The cases from this series were divided into two groups depending on the characteristics of the CT scanners employed. From 1983 to 1985, 78 patients were examined with 3nd-generation CT units; from 1986 to 1989, 63 patients underwent CT performed with 3rd-generation scanner. The results from the two groups were analyzed according to these parameters. A total number of 622 biopsies were performed, of spleen, liver and lymph nodes. CT sensitivity, specificity, and overall accuracy were: 22.9% (group I) vs 43.7% (group II), 83.1% vs 92%, and 68.4% vs 81.2% for lymph nodes; 28.1% vs 36.3%, 93.5% vs 98%, and 66.7% vs 87.3% for the spleen, and 12.5% vs 42.8%, 97.1% vs 98.2% and 88.5% vs 92.1% for the liver. Our results demonstrate an obvious increase in reliability with newer units, even though a high percentage of false-negatives were still observed in our series, which caused understaging in 19.4% of cases vs 24.4 in group I

  14. Diagnostic accuracy of cone-beam CT in the assessment of mandibular invasion of lower gingival carcinoma: Comparison with conventional panoramic radiography

    International Nuclear Information System (INIS)

    Momin, Mohammad A.; Okochi, Kiyoshi; Watanabe, Hiroshi; Imaizumi, Akiko; Omura, Ken; Amagasa, Teruo; Okada, Norihiko; Ohbayashi, Naoto; Kurabayashi, Tohru

    2009-01-01

    Purpose: To evaluate the diagnostic accuracy of cone-beam CT in assessing mandibular invasion by lower gingival carcinoma and compare it with that of panoramic radiography. Patients and methods: Fifty patients with squamous cell carcinoma of the lower gingiva who were examined by both panoramic radiography and cone-beam CT before surgery were included in this study. Five radiologists used a 6-point rating scale to independently evaluate cone-beam CT and panoramic images for the presence or absence of alveolar bone and mandibular canal involvement by tumor. Using the histopathogical findings as the gold standard, we calculated and compared the area under the receiver operating characteristic curve (Az value) and the sensitivity and specificity of the two imaging modalities. Results: In evaluations of both alveolar bone and mandibular canal involvement, the mean Az value for cone-beam CT (0.918 and 0.977, respectively) was significantly higher than that for panoramic radiography (0.793 and 0.872, respectively). The mean sensitivity for cone-beam CT (89% and 99%, respectively) was significantly higher than that for panoramic radiography (73% and 56%, respectively). There was no significant difference in the mean specificity. While cone-beam CT could provide high-resolution three-dimensional images, the image quality around the alveolar crest was often hampered by severe dental artifacts and image noise, resulting in difficulties in detecting subtle alveolar invasion. Conclusion: Cone-beam CT was significantly superior to panoramic radiography in evaluating mandibular invasion by lower gingival carcinoma. Its diagnostic value in detecting subtle alveolar invasion, however, may be limited by severe dental artifacts and image noise.

  15. Impact factor of relationships between CT value and relative electron density for treatment planning system

    International Nuclear Information System (INIS)

    Feng Guosheng; Liang Yuan; Wu Danling; Hao Yanrong; Lu Heming; Chen Jiaxin; Liao Chaolong; Mo Ying; Huang Yihang

    2012-01-01

    Objective: To study the CT values of certain phantoms scanned by various CT scanners with dissimilar parameters. Methods: The CT values of tissue equivalent inserts was measured in the TM164 and CIRS-062 phantom scanned by TOSHIBA AQUILION TM , SIEMENS SOMATOM TM SENSATION TM 64 and SIEMENS SOMATOM TM SENSATION TM OPEN with different voltages,currents and slice thicknesses and then the corresponding CT-to-density curves was compared. Results: There are no significant differences of CT values with various currents and slice thicknesses and also for low atom number materials scanned by different scanners with various tube voltages. The CT values of high atom number materials have obvious differences scanned with tube voltage,the maximum is about 400 HU. There are also significant differences between CT-density curves of two phantoms in the range from soft tissues to dense bone, the maximum is up to 500 HU. Conclusions: CT-density curves were highly affected by materials of phantoms, scanners and tube voltages. It is necessary to measure the curve with a comfortable phantom and certain scanner to assure the accuracy for dose calculation for treatment planning system. (authors)

  16. Means to verify the accuracy of CT systems for metrology applications (In the Absence of Established International Standards)

    International Nuclear Information System (INIS)

    Lettenbauer, H.; Georgi, B.; Weib, D.

    2007-01-01

    X-ray computed tomography (CT) reconstructs an unknown object from X-ray projections and has long been used for qualitative investigation of internal structures in industrial applications. Recently there has been increased interest in applying X-ray cone beam CT to the task of high-precision dimensional measurements of machined parts, since it is a relatively fast method of measuring both inner and outer geometries of arbitrary complexity. The important information for the user in dimensional metrology is if measured elements of a machined part are within the defined tolerances or not. In order to qualify cone beam CT as an established measurement technology, it must be qualified in the same manner as established measurement technologies such as coordinate measurement machines (CMMs) with tactile or optical sensors. In international standards artefacts are defined that are calibrated by certified institutions. These artefacts are defined by certain geometrical elements. CT measurements are performed on the reconstructed object volume, either directly or using an intermediate surface-extraction step. The results of these measurements have to be compared to the values of the calibrated elements; the level of agreement of the results defines the accuracy of the measurements. By using established methods to define measurement uncertainty a very high level of acceptance in dimensional metrology can be reached for the user. Only if results are comparable to standards of the established technologies the barriers of entry into metrology will be removed and all benefits of this technology will be available for the user. (authors)

  17. Accuracy of computer-assisted cervicle pedicle screw installation

    International Nuclear Information System (INIS)

    Zhang Honglei; Zhou Dongsheng; Jang Zhensong

    2009-01-01

    Objective: To investigate the accuracy of computer-assisted cervical pedicle screw installation and the reason of screw malposition. Methods: A total of 172 cervical pedicle screws were installed by computer-assisted navigation for 30 patients with lower cervical spinal diseases. All the patients were examined by X-ray and CT after operation. Screw's position and direction were measured on the sagittal and transectional images of intraoperative navigation and post-operative CT. Then linear regression analysis was taken between navigational and post-operative CT's images. Results: Two screws perforated the upper pedicle wall, 3 perforated the lateral pedicle wall.There was a positive linear correlation between navigational and post-operative CT's images. Conclusion: Computer-assisted navigation can provide the high accuracy of cervical pedicle screw installation and excursion phenomenon is reason of screw malposition. (authors)

  18. Repositioning accuracy of two different mask systems-3D revisited: Comparison using true 3D/3D matching with cone-beam CT

    International Nuclear Information System (INIS)

    Boda-Heggemann, Judit; Walter, Cornelia; Rahn, Angelika; Wertz, Hansjoerg; Loeb, Iris; Lohr, Frank; Wenz, Frederik

    2006-01-01

    Purpose: The repositioning accuracy of mask-based fixation systems has been assessed with two-dimensional/two-dimensional or two-dimensional/three-dimensional (3D) matching. We analyzed the accuracy of commercially available head mask systems, using true 3D/3D matching, with X-ray volume imaging and cone-beam CT. Methods and Materials: Twenty-one patients receiving radiotherapy (intracranial/head-and-neck tumors) were evaluated (14 patients with rigid and 7 with thermoplastic masks). X-ray volume imaging was analyzed online and offline separately for the skull and neck regions. Translation/rotation errors of the target isocenter were analyzed. Four patients were treated to neck sites. For these patients, repositioning was aided by additional body tattoos. A separate analysis of the setup error on the basis of the registration of the cervical vertebra was performed. The residual error after correction and intrafractional motility were calculated. Results: The mean length of the displacement vector for rigid masks was 0.312 ± 0.152 cm (intracranial) and 0.586 ± 0.294 cm (neck). For the thermoplastic masks, the value was 0.472 ± 0.174 cm (intracranial) and 0.726 ± 0.445 cm (neck). Rigid masks with body tattoos had a displacement vector length in the neck region of 0.35 ± 0.197 cm. The intracranial residual error and intrafractional motility after X-ray volume imaging correction for rigid masks was 0.188 ± 0.074 cm, and was 0.134 ± 0.14 cm for thermoplastic masks. Conclusions: The results of our study have demonstrated that rigid masks have a high intracranial repositioning accuracy per se. Given the small residual error and intrafractional movement, thermoplastic masks may also be used for high-precision treatments when combined with cone-beam CT. The neck region repositioning accuracy was worse than the intracranial accuracy in both cases. However, body tattoos and image guidance improved the accuracy. Finally, the combination of both mask systems with 3D

  19. FDG-PET/contrast-enhanced CT as a post-treatment tool in head and neck squamous cell carcinoma: comparison with FDG-PET/non-contrast-enhanced CT and contrast-enhanced CT

    Energy Technology Data Exchange (ETDEWEB)

    Suenaga, Yuko; Kitajima, Kazuhiro; Sugimura, Kazuro [Kobe University Graduate School of Medicine, Department of Radiology, Kobe (Japan); Ishihara, Takeaki; Sasaki, Ryohei [Kobe University Graduate, School of Medicine, Department of Radiology, Division of Radiation Oncology, Kobe (Japan); Otsuki, Naoki; Nibu, Ken-ichi [Kobe University Graduate, School of Medicine, Department of Otolaryngology-Head and Neck Surgery, Kobe (Japan); Minamikawa, Tsutomu [Kobe University Graduate School of Medicine, Department of Oral and Maxillofacial Surgery, Kobe (Japan); Kiyota, Naomi [Kobe University Hospital, Department of Medical Oncology and Hematology, Kobe (Japan)

    2016-04-15

    To evaluate the accuracy of PET/CT using {sup 18}F-fluorodeoxyglucose (FDG) with IV contrast for suspected recurrent head and neck squamous cell carcinoma (HNSCC). One hundred and seventy patients previously treated for HNSCC underwent PET/CT, consisting of non-contrast-enhanced and contrast-enhanced CT, to investigate suspected recurrence. Diagnostic performance of PET/contrast-enhanced CT (PET/ceCT), PET/non-contrast-enhanced CT (PET/ncCT) and contrast-enhanced CT (ceCT) for local or regional recurrence, distant metastasis, overall recurrence and second primary cancer was evaluated. The reference standard included histopathology, treatment change and imaging follow-up. The patient-based areas under the receiver operating characteristic curves (AUC) for ceCT, PET/ncCT and PET/ceCT were 0.82, 0.96 and 0.98 for local recurrence, 0.73, 0.86 and 0.86 for regional recurrence, 0.86, 0.91 and 0.92 for distant metastasis, 0.72, 0.86 and 0.87 for overall recurrence, and 0.86, 0.89 and 0.91 for a second primary cancer. Both PET/ceCT and PET/ncCT statistically showed larger AUC than ceCT for recurrence, and the difference between PET/ceCT and PET/ncCT for local recurrence reached a significant level (p = 0.039). The accuracy of PET/ceCT for diagnosing overall recurrence was high, irrespective of the time interval after the last treatment (83.3-94.1 %). FDG-PET/CT was a more accurate HNSCC restaging tool than ceCT. The added value of ceCT at FDG-PET/CT is minimal. (orig.)

  20. PET/CT和MSCT联合应用在孤立性肺结节诊断中的价值%PET-CT with multi-slice spiral CT in the diagnosis of SPN

    Institute of Scientific and Technical Information of China (English)

    汪世存; 方雷; 潘博; 展凤麟; 谢强; 谢吉奎

    2009-01-01

    目的 探讨18F-脱氧葡萄糖正电子发射显像/电子计算机断层扫描成像(18F-deoxyglucose positron emission tomography /computer tomography 18F-FDG PET/CT)和多排CT( multi-slice spiral CT,MSCT)联合应用在孤立性肺结节(SPN)诊断中的价值.方法 87例SPN患者均行PET/CT及肺部病灶MSCT检查,影像诊断结果 与病理对照,进行统计分析.结果 87例SPN中,MSCT诊断正确64例,误诊23例,误诊率26.4%.PET/CT诊断正确70例,误诊17例,误诊率19.5%.MSCT和PET/CT联合诊断,80例诊断正确,7例误诊,误诊率8.0%.MSCT、PET/CT及两者联合诊断的敏感性、特异性和准确性分别为87.2%、68.1%、79.8%;89.5%、61.3%、81.4%和97.7%、74.2%、91.5%.MSCT和PET/CT单独诊断SPN的准确性无统计学意义(χ2=0.625,P=0.239),MSCT、PET/CT与两者联合应用诊断SPN的准确性均有统计学意义(χ2=7.762和5.318,P=0.005和0.021).结论 SPN的MSCT和PET/CT影像特征均有一定的重叠.单独采用MSCT或PET/CT诊断肺结节无明显统计学差异,MSCT和PET-CT协同诊断的准确性高于MSCT或PET/CT单独诊断.%Aim To evaluate the value of a combination of 18F-FDG PET-CT with multi-slice spiral CT in the diagnosis of SPN.Methods 18F-FDG PET-CT and multi-slice spiral CT were performed in 87 patients with SPN.The pathologically-proved findings were compared with findings diagnosed by MSCT and retrospectively analyzed.Results In 87 cases,the numbers of correct diagnosis and misdiagnosis with MSCT were 64 and 23 respectively,the misdiagnosis rate 26.4%,while the numbers of correct diagnosis and misdiagnosis with PET/CT were 70 and 17,the misdiagnosis rate 19.5%,but 50 cases were correctly diagnosed and 7 cases were misdiagnosed by combining 18F-FDG PET-CT with MSCT,the misdiagnosis rate 8.0%.The sensitivity,specificity,positive predictive value,negative predictive value,and accuracy of MSCT,PET/CT,and MSCT combined with PET /CT were 87.2%,68.1%,79.8%;89.5%,61.3%,81.4% and 97

  1. CT findings in brain edema following the administration of corticosteroids

    International Nuclear Information System (INIS)

    Seki, Yojiro; Kumagai, Norimoto; Aiba, Tadashi

    1979-01-01

    Computed tomography (CT) is the first noninvasive method available for directly visualizing brain edema in man. On CT scans perifocal edema is shown as an area of low density surrounding a lesion. The purpose of this report is to evaluate the effect of corticosteroids on brain edema as seen by CT (HITACHI CT-H 250). Nine patients with brain-tumor and one with brain-abscess were treated with betamethasone for about ten days (dosage started with 12 - 16 mg/day, and tapered). In eight cases, and improvement in the neurological findings was observed. An impressive reduction of peritumoral edema was shown on CT scans in six of these eight cases. There was, however, no significant correlation between the degree of the reduction of edema on CT and that of the improvement in neurological findings. The mode of the CT number in the region of edema did not differ significantly between pre- and post-steroid treatment in the cases showing a recognizable reduction of edema on CT. This failure to change is probably due to the insufficient mechanical accuracy of the CT scanner at the present stage of technology. Through our experiences, it seems that CT is one of the most promising tools for a dynamic study of brain edema in man. (author)

  2. CT findings in brain edema following the administration of corticosteroids

    Energy Technology Data Exchange (ETDEWEB)

    Seki, Y; Kumagai, N; Aiba, T [Toranomon Hospital, Tokyo (Japan)

    1979-03-01

    Computed tomography (CT) is the first noninvasive method available for directly visualizing brain edema in man. On CT scans perifocal edema is shown as an area of low density surrounding a lesion. The purpose of this report is to evaluate the effect of corticosteroids on brain edema as seen by CT (HITACHI CT-H 250). Nine patients with brain-tumor and one with brain-abscess were treated with betamethasone for about ten days (dosage started with 12 - 16 mg/day, and tapered). In eight cases, and improvement in the neurological findings was observed. An impressive reduction of peritumoral edema was shown on CT scans in six of these eight cases. There was, however, no significant correlation between the degree of the reduction of edema on CT and that of the improvement in neurological findings. The mode of the CT number in the region of edema did not differ significantly between pre- and post-steroid treatment in the cases showing a recognizable reduction of edema on CT. This failure to change is probably due to the insufficient mechanical accuracy of the CT scanner at the present stage of technology. Through our experiences, it seems that CT is one of the most promising tools for a dynamic study of brain edema in man.

  3. Accuracy and Utility of Deformable Image Registration in 68Ga 4D PET/CT Assessment of Pulmonary Perfusion Changes During and After Lung Radiation Therapy

    International Nuclear Information System (INIS)

    Hardcastle, Nicholas; Hofman, Michael S.; Hicks, Rodney J.; Callahan, Jason; Kron, Tomas; MacManus, Michael P.; Ball, David L.; Jackson, Price; Siva, Shankar

    2015-01-01

    Purpose: Measuring changes in lung perfusion resulting from radiation therapy dose requires registration of the functional imaging to the radiation therapy treatment planning scan. This study investigates registration accuracy and utility for positron emission tomography (PET)/computed tomography (CT) perfusion imaging in radiation therapy for non–small cell lung cancer. Methods: 68 Ga 4-dimensional PET/CT ventilation-perfusion imaging was performed before, during, and after radiation therapy for 5 patients. Rigid registration and deformable image registration (DIR) using B-splines and Demons algorithms was performed with the CT data to obtain a deformation map between the functional images and planning CT. Contour propagation accuracy and correspondence of anatomic features were used to assess registration accuracy. Wilcoxon signed-rank test was used to determine statistical significance. Changes in lung perfusion resulting from radiation therapy dose were calculated for each registration method for each patient and averaged over all patients. Results: With B-splines/Demons DIR, median distance to agreement between lung contours reduced modestly by 0.9/1.1 mm, 1.3/1.6 mm, and 1.3/1.6 mm for pretreatment, midtreatment, and posttreatment (P<.01 for all), and median Dice score between lung contours improved by 0.04/0.04, 0.05/0.05, and 0.05/0.05 for pretreatment, midtreatment, and posttreatment (P<.001 for all). Distance between anatomic features reduced with DIR by median 2.5 mm and 2.8 for pretreatment and midtreatment time points, respectively (P=.001) and 1.4 mm for posttreatment (P>.2). Poorer posttreatment results were likely caused by posttreatment pneumonitis and tumor regression. Up to 80% standardized uptake value loss in perfusion scans was observed. There was limited change in the loss in lung perfusion between registration methods; however, Demons resulted in larger interpatient variation compared with rigid and B-splines registration. Conclusions

  4. Diagnostic performance of 64-section CT using CT gastrography in preoperative T staging of gastric cancer according to 7th edition of AJCC cancer staging manual

    International Nuclear Information System (INIS)

    Kim, Jin Woong; Shin, Sang Soo; Heo, Suk Hee; Lim, Hyo Soon; Jeong, Yong Yeon; Kang, Heoung Keun; Choi, Yoo Duk; Park, Young Kyu; Park, Chang Hwan

    2012-01-01

    To evaluate the accuracy of 64-section multidetector CT with CT gastrography for determining the depth of mural invasion in patients with gastric cancer according to the 7th edition of the AJCC cancer staging manual. A total of 127 patients with gastric cancer and who had undergone both esophago-gastro-duodenoscopy and 64-section CT were included in this study. Two radiologists independently reviewed the preoperative CT images with respect to the detectability and T-staging of the gastric cancers. The sensitivity, specificity, accuracy and overall accuracy of each reviewer for the T staging of gastric cancer were calculated. Overall, gastric cancer was detected in 123 (96.9%) of the 127 cancers on the CT images. Reviewer 1 correctly staged 98 gastric cancers, and reviewer 2 correctly classified 105 gastric cancers. The overall diagnostic accuracy of the T staging was 77.2% (98/127) for reviewer 1 and 82.7% (105/127) for reviewer 2. 64-section CT using CT gastrography showed a reasonable diagnostic performance for determining the T staging in patients with gastric cancer according to the 7th edition of the AJCC cancer staging manual. (orig.)

  5. Influence of Co-57 and CT Transmission Measurements on the Quantification Accuracy and Partial Volume Effect of a Small Animal PET Scanner.

    Science.gov (United States)

    Mannheim, Julia G; Schmid, Andreas M; Pichler, Bernd J

    2017-12-01

    Non-invasive in vivo positron emission tomography (PET) provides high detection sensitivity in the nano- to picomolar range and in addition to other advantages, the possibility to absolutely quantify the acquired data. The present study focuses on the comparison of transmission data acquired with an X-ray computed tomography (CT) scanner or a Co-57 source for the Inveon small animal PET scanner (Siemens Healthcare, Knoxville, TN, USA), as well as determines their influences on the quantification accuracy and partial volume effect (PVE). A special focus included the impact of the performed calibration on the quantification accuracy. Phantom measurements were carried out to determine the quantification accuracy, the influence of the object size on the quantification, and the PVE for different sphere sizes, along the field of view and for different contrast ratios. An influence of the emission activity on the Co-57 transmission measurements was discovered (deviations up to 24.06 % measured to true activity), whereas no influence of the emission activity on the CT attenuation correction was identified (deviations influenced by the applied calibration factor and by the object size. The PVE demonstrated a dependency on the sphere size, the position within the field of view, the reconstruction and correction algorithms and the count statistics. Depending on the reconstruction algorithm, only ∼30-40 % of the true activity within a small sphere could be resolved. The iterative 3D reconstruction algorithms uncovered substantially increased recovery values compared to the analytical and 2D iterative reconstruction algorithms (up to 70.46 % and 80.82 % recovery for the smallest and largest sphere using iterative 3D reconstruction algorithms). The transmission measurement (CT or Co-57 source) to correct for attenuation did not severely influence the PVE. The analysis of the quantification accuracy and the PVE revealed an influence of the object size, the reconstruction

  6. Spiral CT scanning plan to generate accurate Fe models of the human femur

    International Nuclear Information System (INIS)

    Zannoni, C.; Testi, D.; Capello, A.

    1999-01-01

    In spiral computed tomography (CT), source rotation, patient translation, and data acquisition are continuously conducted. Settings of the detector collimation and the table increment affect the image quality in terms of spatial and contrast resolution. This study assessed and measured the efficacy of spiral CT in those applications where the accurate reconstruction of bone morphology is critical: custom made prosthesis design or three dimensional modelling of the mechanical behaviour of long bones. Results show that conventional CT grants the highest accuracy. Spiral CT with D=5 mm and P=1,5 in the regions where the morphology is more regular, slightly degrades the image quality but allows to acquire at comparable cost an higher number of images increasing the longitudinal resolution of the acquired data set. (author)

  7. A new irradiation unit constructed of self-moving gantry-CT and linac

    International Nuclear Information System (INIS)

    Kuriyama, Kengo; Onishi, Hiroshi; Sano, Naoki; Komiyama, Takafumi; Aikawa, Yoshihito; Tateda, Yoshihito; Araki, Tsutomu; Uematsu, Minoru

    2003-01-01

    Purpose: To improve reproducibility in stereotactic irradiation (STI) without using noninvasive immobilization devices or body frames, we have developed an integrated computed tomography (CT)-linac irradiation system connecting CT scanner and linac via a common treatment couch. Methods and Materials: This system consists of a linac, a CT scanner, and a common treatment couch. The linac and the CT gantry are positioned on opposite ends of the couch so that, by rotating the treatment couch, linac radiotherapy or CT scanning can be performed. The rotational axis of the linac gantry is coaxial with that of the CT gantry, and the position of the linac isocenter on the couch matches the origin of the coordinate system for CT scanning when the couch is rotated 180 degree sign toward the CT side. Instead of the couch moving into the gantry, as in conventional CT, in this case the table is fixed and scanning is accomplished by moving the gantry. We evaluated the rotational accuracy of the common couch and the scan-position accuracy of the self-moving gantry CT. Results: The positional accuracy of the common couch was 0.20, 0.18, and 0.39 mm in the lateral, longitudinal, and vertical directions, respectively. The scan-position accuracy of the CT gantry was less than 0.4 mm in the lateral, longitudinal, and vertical directions. Conclusion: This irradiation system has a high accuracy and is useful for noninvasive STI and for verification of the position of a target in three-dimensional conformal radiotherapy

  8. A quality assurance phantom for the performance evaluation of volumetric micro-CT systems

    Energy Technology Data Exchange (ETDEWEB)

    Du, Louise Y [Department of Medical Biophysics, University of Western Ontario, London, ON (Canada); Umoh, Joseph [Imaging Research Laboratories, Robarts Research Institute, London, ON (Canada); Nikolov, Hristo N [Imaging Research Laboratories, Robarts Research Institute, London, ON (Canada); Pollmann, Steven I [Imaging Research Laboratories, Robarts Research Institute, London, ON (Canada); Lee, Ting-Yim [Department of Medical Biophysics, University of Western Ontario, London, ON (Canada); Holdsworth, David W [Department of Medical Biophysics, University of Western Ontario, London, ON (Canada)

    2007-12-07

    Small-animal imaging has recently become an area of increased interest because more human diseases can be modeled in transgenic and knockout rodents. As a result, micro-computed tomography (micro-CT) systems are becoming more common in research laboratories, due to their ability to achieve spatial resolution as high as 10 {mu}m, giving highly detailed anatomical information. Most recently, a volumetric cone-beam micro-CT system using a flat-panel detector (eXplore Ultra, GE Healthcare, London, ON) has been developed that combines the high resolution of micro-CT and the fast scanning speed of clinical CT, so that dynamic perfusion imaging can be performed in mice and rats, providing functional physiological information in addition to anatomical information. This and other commercially available micro-CT systems all promise to deliver precise and accurate high-resolution measurements in small animals. However, no comprehensive quality assurance phantom has been developed to evaluate the performance of these micro-CT systems on a routine basis. We have designed and fabricated a single comprehensive device for the purpose of performance evaluation of micro-CT systems. This quality assurance phantom was applied to assess multiple image-quality parameters of a current flat-panel cone-beam micro-CT system accurately and quantitatively, in terms of spatial resolution, geometric accuracy, CT number accuracy, linearity, noise and image uniformity. Our investigations show that 3D images can be obtained with a limiting spatial resolution of 2.5 mm{sup -1} and noise of {+-}35 HU, using an acquisition interval of 8 s at an entrance dose of 6.4 cGy.

  9. A quality assurance phantom for the performance evaluation of volumetric micro-CT systems

    International Nuclear Information System (INIS)

    Du, Louise Y; Umoh, Joseph; Nikolov, Hristo N; Pollmann, Steven I; Lee, Ting-Yim; Holdsworth, David W

    2007-01-01

    Small-animal imaging has recently become an area of increased interest because more human diseases can be modeled in transgenic and knockout rodents. As a result, micro-computed tomography (micro-CT) systems are becoming more common in research laboratories, due to their ability to achieve spatial resolution as high as 10 μm, giving highly detailed anatomical information. Most recently, a volumetric cone-beam micro-CT system using a flat-panel detector (eXplore Ultra, GE Healthcare, London, ON) has been developed that combines the high resolution of micro-CT and the fast scanning speed of clinical CT, so that dynamic perfusion imaging can be performed in mice and rats, providing functional physiological information in addition to anatomical information. This and other commercially available micro-CT systems all promise to deliver precise and accurate high-resolution measurements in small animals. However, no comprehensive quality assurance phantom has been developed to evaluate the performance of these micro-CT systems on a routine basis. We have designed and fabricated a single comprehensive device for the purpose of performance evaluation of micro-CT systems. This quality assurance phantom was applied to assess multiple image-quality parameters of a current flat-panel cone-beam micro-CT system accurately and quantitatively, in terms of spatial resolution, geometric accuracy, CT number accuracy, linearity, noise and image uniformity. Our investigations show that 3D images can be obtained with a limiting spatial resolution of 2.5 mm -1 and noise of ±35 HU, using an acquisition interval of 8 s at an entrance dose of 6.4 cGy

  10. Improving the false-negative rate of CT in acute appendicitis-Reassessment of CT images by body imaging radiologists: A blinded prospective study

    International Nuclear Information System (INIS)

    Poortman, Pieter; Lohle, Paul N.M.; Schoemaker, Cees M.; Cuesta, Miguel A.; Oostvogel, Henk J.M.; Lange-de Klerk, Elly S.M. de; Hamming, Jaap F.

    2010-01-01

    Purpose: To compare the accuracy of computed tomography (CT) analyzed by individual radiology staff members and body imaging radiologists in a non-academic teaching hospital for the diagnosis of acute appendicitis. Patients and methods: In a prospective study 199 patients with suspected acute appendicitis were examined with unenhanced CT. CT images were pre-operatively analyzed by one of the 12 members of the radiology staff. In a later stage two body imaging radiologist reassessed all CT images without knowledge of the surgical findings and without knowledge of the primary CT diagnosis. The results, independently reported, were correlated with surgical and histopathologic findings. Results: In 132 patients (66%) acute appendicitis was found at surgery, in 67 patients (34%) a normal appendix was found. The sensitivity of the primary CT analysis and of the reassessment was 76% and 88%, respectively; the specificity was 84% and 87%; the positive predictive value was 90% and 93%; the negative predictive value was 64% and 78%; and the accuracy was 78% and 87%. Conclusion: Reassessment of CT images for acute appendicitis by body imaging radiologists results in a significant improvement of sensitivity, negative predictive value and accuracy. To prevent false-negative interpretation of CT images in acute appendicitis the expertise of the attending radiologist should be considered.

  11. Improving the false-negative rate of CT in acute appendicitis-Reassessment of CT images by body imaging radiologists: A blinded prospective study

    Energy Technology Data Exchange (ETDEWEB)

    Poortman, Pieter [Department of Surgery, St Elisabeth Hospital, Tilburg (Netherlands)], E-mail: ppoortman@wlz.nl; Lohle, Paul N.M. [Department of Surgery, St Elisabeth Hospital, Tilburg (Netherlands)], E-mail: plohle@elisabeth.nl; Schoemaker, Cees M. [Department of Surgery, St Elisabeth Hospital, Tilburg (Netherlands)], E-mail: mcschoemaker@elisabeth.nl; Cuesta, Miguel A. [Department of Surgery, VU Medical Centre, Amsterdam (Netherlands)], E-mail: ma.cuesta@vumc.nl; Oostvogel, Henk J.M. [Department of Surgery, St Elisabeth Hospital, Tilburg (Netherlands)], E-mail: h.oostvogel@elisabeth.nl; Lange-de Klerk, Elly S.M. de [Department of Epidemiology and Biostatistics, VU Medical Centre, Amsterdam (Netherlands)], E-mail: esm.delange@vumc.nl; Hamming, Jaap F. [Department of Surgery, Leiden University Medical Centre (Netherlands)], E-mail: j.f.hamming@lumc.nl

    2010-04-15

    Purpose: To compare the accuracy of computed tomography (CT) analyzed by individual radiology staff members and body imaging radiologists in a non-academic teaching hospital for the diagnosis of acute appendicitis. Patients and methods: In a prospective study 199 patients with suspected acute appendicitis were examined with unenhanced CT. CT images were pre-operatively analyzed by one of the 12 members of the radiology staff. In a later stage two body imaging radiologist reassessed all CT images without knowledge of the surgical findings and without knowledge of the primary CT diagnosis. The results, independently reported, were correlated with surgical and histopathologic findings. Results: In 132 patients (66%) acute appendicitis was found at surgery, in 67 patients (34%) a normal appendix was found. The sensitivity of the primary CT analysis and of the reassessment was 76% and 88%, respectively; the specificity was 84% and 87%; the positive predictive value was 90% and 93%; the negative predictive value was 64% and 78%; and the accuracy was 78% and 87%. Conclusion: Reassessment of CT images for acute appendicitis by body imaging radiologists results in a significant improvement of sensitivity, negative predictive value and accuracy. To prevent false-negative interpretation of CT images in acute appendicitis the expertise of the attending radiologist should be considered.

  12. Accuracy Enhancement of CT Measurements using Data Filtering

    DEFF Research Database (Denmark)

    Stolfi, Alessandro; Kallasse, Maarja-Helena; Carli, Lorenzo

    2016-01-01

    This paper describes the impact of data filtering on CT capability to inspect assemblies. The investigation was carried out using an industrial multi-material assembly provided by Novo Nordisk A/S. The assembly comprises two parts made of polyoxymethylene (POM) and of an alloy comprising polycarb...

  13. Understanding less than nothing: Children’s neural response to negative numbers shifts across age and accuracy

    Directory of Open Access Journals (Sweden)

    Margaret M Gullick

    2013-09-01

    Full Text Available We examined the brain activity underlying the development of our understanding of negative numbers, which are amounts lacking direct physical counterparts. Children performed a paired comparison task with positive and negative numbers during an fMRI session. As previously shown in adults, both pre-instruction fifth graders and post-instruction seventh graders demonstrated typical behavioral and neural distance effects to negative numbers, where response times and parietal and frontal activity increased as comparison distance decreased. We then determined the factors impacting the distance effect in each age group. Behaviorally, the fifth grader distance effect for negatives was significantly predicted only by positive comparison accuracy, indicating that children who were generally better at working with numbers were better at comparing negatives. In seventh graders, negative comparison accuracy significantly predicted their distance effect, indicating that children who were better at working with negative numbers demonstrated a more typical distance effect. Across children, as age increased, the negative number distance effect increased in the bilateral IPS and decreased frontally, indicating a frontoparietal shift consistent with previous numerical development literature. In contrast, as task accuracy increased, the parietal distance effect increased in the left IPS and decreased in the right, possibly indicating a change from an approximate understanding of negatives’ values to a more exact, precise representation (particularly supported by the left IPS with increasing expertise. These shifts separately indicate the effects of increasing maturity generally in numeric processing and specifically in negative number understanding.

  14. Diagnostic accuracy of 128-slice dual-source CT coronary angiography: a randomized comparison of different acquisition protocols

    International Nuclear Information System (INIS)

    Neefjes, Lisan A.; Kate, Gert-Jan R. ten; Rossi, Alexia; Nieman, Koen; Papadopoulou, Stella L.; Dharampal, Anoeshka S.; Dedic, Admir; Feyter, Pim J. de; Mollet, Nico R.; Genders, Tessa S.S.; Hunink, M.G.M.; Schultz, Carl J.; Weustink, Annick C.; Dijkshoorn, Marcel L.; Straten, Marcel van; Cademartiri, Filippo; Krestin, Gabriel P.

    2013-01-01

    To compare the diagnostic performance and radiation exposure of 128-slice dual-source CT coronary angiography (CTCA) protocols to detect coronary stenosis with more than 50 % lumen obstruction. We prospectively included 459 symptomatic patients referred for CTCA. Patients were randomized between high-pitch spiral vs. narrow-window sequential CTCA protocols (heart rate below 65 bpm, group A), or between wide-window sequential vs. retrospective spiral protocols (heart rate above 65 bpm, group B). Diagnostic performance of CTCA was compared with quantitative coronary angiography in 267 patients. In group A (231 patients, 146 men, mean heart rate 58 ± 7 bpm), high-pitch spiral CTCA yielded a lower per-segment sensitivity compared to sequential CTCA (89 % vs. 97 %, P = 0.01). Specificity, PPV and NPV were comparable (95 %, 62 %, 99 % vs. 96 %, 73 %, 100 %, P > 0.05) but radiation dose was lower (1.16 ± 0.60 vs. 3.82 ± 1.65 mSv, P 0.05). Radiation dose of sequential CTCA was lower compared to retrospective CTCA (6.12 ± 2.58 vs. 8.13 ± 4.52 mSv, P < 0.001). Diagnostic performance was comparable in both groups. Sequential CTCA should be used in patients with regular heart rates using 128-slice dual-source CT, providing optimal diagnostic accuracy with as low as reasonably achievable (ALARA) radiation dose. circle 128-slice dual-source CT coronary angiography offers several different acquisition protocols. (orig.)

  15. Post-operative 3D CT feedback improves accuracy and precision in the learning curve of anatomic ACL femoral tunnel placement.

    Science.gov (United States)

    Sirleo, Luigi; Innocenti, Massimo; Innocenti, Matteo; Civinini, Roberto; Carulli, Christian; Matassi, Fabrizio

    2018-02-01

    To evaluate the feedback from post-operative three-dimensional computed tomography (3D-CT) on femoral tunnel placement in the learning process, to obtain an anatomic anterior cruciate ligament (ACL) reconstruction. A series of 60 consecutive patients undergoing primary ACL reconstruction using autologous hamstrings single-bundle outside-in technique were prospectively included in the study. ACL reconstructions were performed by the same trainee-surgeon during his learning phase of anatomic ACL femoral tunnel placement. A CT scan with dedicated tunnel study was performed in all patients within 48 h after surgery. The data obtained from the CT scan were processed into a three-dimensional surface model, and a true medial view of the lateral femoral condyle was used for the femoral tunnel placement analysis. Two independent examiners analysed the tunnel placements. The centre of femoral tunnel was measured using a quadrant method as described by Bernard and Hertel. The coordinates measured were compared with anatomic coordinates values described in the literature [deep-to-shallow distance (X-axis) 28.5%; high-to-low distance (Y-axis) 35.2%]. Tunnel placement was evaluated in terms of accuracy and precision. After each ACL reconstruction, results were shown to the surgeon to receive an instant feedback in order to achieve accurate correction and improve tunnel placement for the next surgery. Complications and arthroscopic time were also recorded. Results were divided into three consecutive series (1, 2, 3) of 20 patients each. A trend to placing femoral tunnel slightly shallow in deep-to-shallow distance and slightly high in high-to-low distance was observed in the first and the second series. A progressive improvement in tunnel position was recorded from the first to second series and from the second to the third series. Both accuracy (+52.4%) and precision (+55.7%) increased from the first to the third series (p process to improve accuracy and precision of femoral

  16. Does Antibiotic Treatment Affect the Diagnostic Accuracy of 18F-FDG PET/CT Studies in Patients with Suspected Infectious Processes?

    Science.gov (United States)

    Kagna, Olga; Kurash, Marina; Ghanem-Zoubi, Nesrin; Keidar, Zohar; Israel, Ora

    2017-11-01

    18 F-FDG PET/CT plays a significant role in the assessment of various infectious processes. Patients with suspected or known sites of infection are often referred for 18 F-FDG imaging while already receiving antibiotic treatment. The current study assessed whether antibiotic therapy affected the detectability rate of infectious processes by 18 F-FDG PET/CT. Methods: A 5-y retrospective study of all adult patients who underwent 18 F-FDG PET/CT in search of a focal source of infection was performed. The presence, duration, and appropriateness of antibiotic treatment before 18 F-FDG imaging were recorded. Diagnosis of an infectious process was based on microbiologic or pathologic data as well as on clinical and radiologic follow-up. Results: Two hundred seventeen patients underwent 243 PET/CT studies in search of a focal source of infection and were included in the study. Sixty-seven studies were excluded from further analysis because of a final noninfectious etiology or lack of further follow-up or details regarding the antibiotic treatment. The final study population included 176 18 F-FDG PET/CT studies in 153 patients (107 men, 46 women; age range, 18-86 y). One hundred nineteen studies (68%) were performed in patients receiving antibiotic therapy for a range of 1-73 d. A diagnosis of infection was made in 107 true-positive cases (61%), including 63 studies (59%) in patients receiving appropriate antibiotic therapy started before the performance of the 18 F-FDG PET/CT study. There were 52 true-negative (29%) and 17 false-positive (10%) 18 F-FDG PET/CT studies. No false-negative results were found. Conclusion: 18 F-FDG PET/CT correctly identified foci of increased uptake compatible with infection in most patients, including all patients receiving appropriate antimicrobial therapy, with no false-negative cases. On the basis of the current study results, the administration of antibiotics appears to have no clinically significant impact on the diagnostic accuracy of 18

  17. TH-CD-206-08: An Anthropopathic Deformable Phantom for Geometric and Dose Accumulation Accuracy Validation of Deformable Image Registration

    Energy Technology Data Exchange (ETDEWEB)

    Liao, Y; Chen, H; Chen, J; Zhen, X; Zhou, L [Southern Medical University, Guangzhou, Guangdong (China); Gu, X [UT Southwestern Medical Center, Dallas, TX (United States)

    2016-06-15

    Purpose: To design and construct a three-dimensional (3D) anthropopathic abdominal phantom for evaluating deformable image registration (DIR) accuracy on images and dose deformation in adaptive radiation therapy (ART). Method: Organ moulds, including liver, kidney, spleen, stomach, vertebra and two metastasis tumors, are 3D printed using the contours from an ovarian cancer patient. The organ moulds are molded with deformable gels that made of different mixtures of polyvinyl chloride (PVC) and the softener dioctyl terephthalate. Gels with different densities are obtained by a polynomial fitting curve which describes the relation between the CT number and PVC-softener blending ratio. The rigid vertebras are constructed by moulding with white cement. The final abdominal phantom is assembled by arranging all the fabricated organs inside a hollow dummy according to their anatomies and sealed with deformable gel with averaged CT number of muscle and fat. Geometric and dosimetric landmarks are embedded inside the phantom for spatial accuracy and dose accumulation accuracy studies. Three DIR algorithms available in the open source DIR toolkit-DIRART, including the Demons, the Horn-Schunck and Lucas-Kanade method and the Level-Set Motion method, are tested using the constructed phantom. Results: Viscoelastic behavior is observed in the constructed deformable gel, which serves as an ideal material for the deformable phantom. The constructed abdominal phantom consists of highly realistic anatomy and the fabricated organs inside have close CT number to its reference patient. DIR accuracy studies conducted on the constructed phantom using three DIR approaches indicate that geometric accuracy of a DIR algorithm has achieved does not guarantee accuracy in dose accumulation. Conclusions: We have designed and constructed an anthropopathic abdominal deformable phantom with satisfactory elastic property, realistic organ density and anatomy. This physical phantom is recyclable and can

  18. Helical CT imaging of clinically suspected appendicitis: Correlation of CT and histological findings

    International Nuclear Information System (INIS)

    Wong, S.K.; Chan, L.P.; Yeo, A.

    2002-01-01

    PURPOSE: The diagnosis of appendicitis is traditionally made on the basis of clinical findings supported by laboratory results. The aim of our study was to determine the accuracy and feasibility of using a relatively new technique of computed tomography (CT) using only colonic contrast medium. MATERIALS AND METHODS: A total of 50 patients clinically diagnosed as having appendicitis were prospectively examined before surgery with thin-collimation helical CT from the L3 level to the acetabular roof with only rectally administered colon contrast medium. The hard copy CT images were reviewed jointly by two radiologists and a consensus was reached for each patient. The results were then compared with the surgical and histological findings at appendicectomy. RESULTS: There were 35 true-positives, one false-positive, 12 true-negatives and two false-negatives for CT. This yielded an accuracy of 94%, sensitivity of 95%, specificity of 92%, positive predictive value of 97% and negative predictive value of 86%. The appendix was identified in 45 patients (90%) and obscured by an inflammatory mass in the remaining five. An alternative diagnosis was found in 10 of 12 normal CT examinations (83%). CONCLUSION: Helical CT with rectal contrast medium is a quick, well tolerated and accurate test to diagnose appendicitis. It can offer alternative, possibly non-surgical diagnosis in patients who would otherwise have undergone laparotomy. Wong, S.K. et al. (2002)

  19. Three-dimensional CT pyelography for planning of percutaneous nephrostolithotomy: accuracy of stone measurement, stone depiction and pelvicalyceal reconstruction

    Energy Technology Data Exchange (ETDEWEB)

    Patel, Uday [St George' s Hospital and Medical School, Department of Radiology, London (United Kingdom); Princess Grace Hospital, London (United Kingdom); Walkden, Richard Miles [St George' s Hospital and Medical School, Department of Radiology, London (United Kingdom); Ghani, Khurshid R. [St George' s Hospital and Medical School, Department of Urology, London (United Kingdom); Anson, Ken [St George' s Hospital and Medical School, Department of Urology, London (United Kingdom); Princess Grace Hospital, London (United Kingdom)

    2009-05-15

    Retrospective evaluation of computed tomographic (CT) pyelography before percutaneous nephrostolithotomy (PCNL). Twenty patients with renal calculi underwent CT pyelography using a dedicated protocol. Calculus size, uniformity of contrast excretion and accuracy of calculus and pelvicalyceal (PC) system reconstructions were scored and compared on axial and coronal maximum intensity projections (MIP) and volume reconstructions (VRmovie loops). After contrast medium administration, the size of calculi is accurate on axial images, but underestimated on coronal studies: mean 14.7 mm vs. 14.4 mm (axial) and 17.2 mm vs. 16.1 mm (coronal) for measurements before and after enhancement, respectively (p = 0.11 and 0.03). Uniform contrast medium excretion (median 228 HU; 95% CI 209-266 HU) was sufficiently lower than calculus density (median 845 HU; 95% CI 457-1,193 HU) for precise calculus and pelvicalyceal reconstructions in 87% and 85%, respectively. Coronal MIP scans were rated best for calculus depiction (mean score 2.68 vs. 2.50 and 2.41 for coronal, axial and VRs, respectively; p = 0.14) and VR studies best for PC anatomy (mean score 4.4 vs. 3.73 and 2.89 for VR, coronal and axial studies, respectively; p = <0.0001). Three-dimensional CT pyelography can accurately demonstrate calculus position and spatial relationships of the collecting system before PCNL. (orig.)

  20. Three-dimensional CT pyelography for planning of percutaneous nephrostolithotomy: accuracy of stone measurement, stone depiction and pelvicalyceal reconstruction

    International Nuclear Information System (INIS)

    Patel, Uday; Walkden, Richard Miles; Ghani, Khurshid R.; Anson, Ken

    2009-01-01

    Retrospective evaluation of computed tomographic (CT) pyelography before percutaneous nephrostolithotomy (PCNL). Twenty patients with renal calculi underwent CT pyelography using a dedicated protocol. Calculus size, uniformity of contrast excretion and accuracy of calculus and pelvicalyceal (PC) system reconstructions were scored and compared on axial and coronal maximum intensity projections (MIP) and volume reconstructions (VRmovie loops). After contrast medium administration, the size of calculi is accurate on axial images, but underestimated on coronal studies: mean 14.7 mm vs. 14.4 mm (axial) and 17.2 mm vs. 16.1 mm (coronal) for measurements before and after enhancement, respectively (p = 0.11 and 0.03). Uniform contrast medium excretion (median 228 HU; 95% CI 209-266 HU) was sufficiently lower than calculus density (median 845 HU; 95% CI 457-1,193 HU) for precise calculus and pelvicalyceal reconstructions in 87% and 85%, respectively. Coronal MIP scans were rated best for calculus depiction (mean score 2.68 vs. 2.50 and 2.41 for coronal, axial and VRs, respectively; p = 0.14) and VR studies best for PC anatomy (mean score 4.4 vs. 3.73 and 2.89 for VR, coronal and axial studies, respectively; p = <0.0001). Three-dimensional CT pyelography can accurately demonstrate calculus position and spatial relationships of the collecting system before PCNL. (orig.)

  1. {sup 18}F-Fluorocholine PET/CT as a complementary tool in the follow-up of low-grade glioma: diagnostic accuracy and clinical utility

    Energy Technology Data Exchange (ETDEWEB)

    Gomez-Rio, Manuel; Rodriguez-Fernandez, Antonio; Llamas-Elvira, Jose M. [Instituto de Investigacion Biosanitaria de Granada, Granada (Spain); University Hospital ' ' Virgen de las Nieves' ' , Department of Nuclear Medicine, Granada (Spain); Testart Dardel, Nathalie [University Hospital ' ' Virgen de las Nieves' ' , Department of Nuclear Medicine, Granada (Spain); Santiago Chinchilla, Alicia [University Hospital ' ' Virgen de las Nieves' ' , Department of Radiology, Granada (Spain); Olivares Granados, Gonzalo [University Hospital ' ' Virgen de las Nieves' ' , Department of Neurosurgery, Granada (Spain); Luque Caro, Raquel [University Hospital ' ' Virgen de las Nieves' ' , Department of Medical Oncology, Granada (Spain); Zurita Herrera, Mercedes [University Hospital ' ' Virgen de las Nieves' ' , Department of Radiotherapy Oncology, Granada (Spain); Chamorro Santos, Clara E. [University Hospital ' ' Virgen de las Nieves' ' , Department of Pathology, Granada (Spain); Lardelli-Claret, Pablo [Instituto de Investigacion Biosanitaria de Granada, Granada (Spain); Centros de Investigacion Biomedica en Red de Epidemiologia y Salud Publica (CIBERESP), Granada (Spain); University of Granada, Department of Preventive Medicine and Public Health, School of Medicine, Granada (Spain)

    2015-05-01

    The follow-up of treated low-grade glioma (LGG) requires the evaluation of subtle clinical changes and MRI results. When the result is inconclusive, additional procedures are required to assist decision-making, such as the use of advanced MRI (aMRI) sequences and nuclear medicine scans (SPECT and PET). The aim of this study was to determine whether incorporating {sup 18}F-fluorocholine PET/CT in the follow-up protocol for treated LGG improves diagnostic accuracy and clinical utility. This was a prospective case-series study in patients with treated LGG during standard follow-up with indeterminate clinical and/or radiological findings of tumour activity. All patients underwent clinical evaluation, aMRI, {sup 201}Tl-SPECT and {sup 18}F-fluorocholine PET/CT. Images were interpreted by visual evaluation complemented with semiquantitative analysis. Between January 2012 and December 2013, 18 patients were included in this study. The final diagnosis was established by histology (five surgical specimens, one biopsy specimen) or by consensus of the Neuro-Oncology Group (11 patients) after a follow-up of >6 months (mean 14.9 ± 2.72 months). The global diagnostic accuracies were 90.9 % for aMRI (38.8 % inconclusive), 69.2 % for {sup 201}Tl-SPECT (11.1 % inconclusive), and 100 % for {sup 18}F-fluorocholine PET/CT. {sup 201}Tl-SPECT led correctly to a change in the initial approach in 38.9 % of patients but might have led to error in 27.8 %. The use of {sup 18}F-fluorocholine PET/CT alone rather than {sup 201}Tl-SPECT led correctly to a change in the approach suggested by routine follow-up in 72.2 % of patients and endorsed the approach in the remaining 27.8 %. Our results support the need to complement structural MRI with aMRI and nuclear medicine procedures in selected patients. {sup 18}F-Fluorocholine PET/CT can be useful in the individualized management of patients with treated LGG with uncertain clinical and/or radiological evidence of tumour activity. (orig.)

  2. Machine learning-based dual-energy CT parametric mapping.

    Science.gov (United States)

    Su, Kuan-Hao; Kuo, Jung-Wen; Jordan, David W; Van Hedent, Steven; Klahr, Paul; Wei, Zhouping; Al Helo, Rose; Liang, Fan; Qian, Pengjiang; Pereira, Gisele C; Rassouli, Negin; Gilkeson, Robert C; Traughber, Bryan J; Cheng, Chee-Wai; Muzic, Raymond F

    2018-05-22

    The aim is to develop and evaluate machine learning methods for generating quantitative parametric maps of effective atomic number (Zeff), relative electron density (ρe), mean excitation energy (Ix), and relative stopping power (RSP) from clinical dual-energy CT data. The maps could be used for material identification and radiation dose calculation. Machine learning methods of historical centroid (HC), random forest (RF), and artificial neural networks (ANN) were used to learn the relationship between dual-energy CT input data and ideal output parametric maps calculated for phantoms from the known compositions of 13 tissue substitutes. After training and model selection steps, the machine learning predictors were used to generate parametric maps from independent phantom and patient input data. Precision and accuracy were evaluated using the ideal maps. This process was repeated for a range of exposure doses, and performance was compared to that of the clinically-used dual-energy, physics-based method which served as the reference. The machine learning methods generated more accurate and precise parametric maps than those obtained using the reference method. Their performance advantage was particularly evident when using data from the lowest exposure, one-fifth of a typical clinical abdomen CT acquisition. The RF method achieved the greatest accuracy. In comparison, the ANN method was only 1% less accurate but had much better computational efficiency than RF, being able to produce parametric maps in 15 seconds. Machine learning methods outperformed the reference method in terms of accuracy and noise tolerance when generating parametric maps, encouraging further exploration of the techniques. Among the methods we evaluated, ANN is the most suitable for clinical use due to its combination of accuracy, excellent low-noise performance, and computational efficiency. . © 2018 Institute of Physics and Engineering in

  3. Role of CT scan in diagnosis of acute appendicitis

    International Nuclear Information System (INIS)

    Salem, O.A.; Khasawneh, M.

    2007-01-01

    To evaluate the diagnostic accuracy of the spiral-CT in patients with clinically suspected acute appendicitis. A total of 124 patients referred for CT scan with suspected appendicitis between January 2005 and October 2006 were assessed for the appendiceal size and the presence of signs of appendicitis. The findings were correlated with surgical histopathology. CT scan had a sensitivity of 95 percent and a specifity of 93 percent and an overall accuracy of 92 percent. The use of spiral CT in patients with equivocal clinical presentation suspected of having acute appendicitis can lead to significant improvement in the preoperative diagnosis. (author)

  4. Accuracy of preoperative diagnosis for primary hyperparathyroidism by CT, US and scintigraphy

    International Nuclear Information System (INIS)

    Imada, Masanobu; Nonaka, Satoshi; Hayashi, Tatsuya; Kunibe, Isamu; Harabuchi, Yasuaki

    2003-01-01

    We determined the usefulness of computed tomography (CT), ultrasonography (US) and Tc-Tl subtraction scintigraphy for preoperative diagnosis in 12 patients with primary hyperparathyroidism, all of which were solitary adenoma. Seven (63.6%) of 11 tumors were detected by CT, 7 (58.3%) of 12 tumors were detected by US, and 7 (63.6%) of 11 tumors were detected by Tc-Tl subtraction scintigraphy. Of 7 tumors with size>10 mm, 6 (85.7%) were detected by CT, 5 (71.4%) were detected by US, and 6 (85.7%) were detected by Tc-Tl subtraction scintigraphy, whereas only 1 (25%) with a diameter of 10 mm was detected by each of CT or Tc-Tl subtraction scintigraphy and 2 (40%) with diameters 10 mm were detected by US. Of 6 tumors situated behind the thyroid gland or intra thyroid gland, 5 (83.3%) were detected by CT and all 6 (100%) were detected by US, Tl-Tc subtraction scintigraphy, which was employed for 5 tumors, detected all 5 (100%). However, in the tumors situated beneath the thyroid gland only 2 (40%) of 5, 1 (16.7%) of 6 and 2 (33.3%) of 6 were detected by CT, US and Tc-Tl subtraction scintigraphy respectively. In 5 tumors without thyroid disease, 4 (80%) of 5 were detected by both CT and US, 3 (75%) of 4 were detected by Tc-Tl subtraction scintigraphy. However, in 7 tumors with thyroid diseases (papillary carcinoma, follicular carcinoma, follicular adenoma, chronic thyroiditis, and adenomatous goiter), 3 (50%) of 6 were detected by CT, 3 (42.9%) of 7 were detected by US and 4 (57.1%) of 7 were detected by Tc-Tl subtraction scintigraphy. Location of tumor was the factor most likely to make appropriate diagnosis difficult for primary hyperparathyroidism by any assessment method. (author)

  5. Optimizing imaging in suspected appendicitis (OPTIMAP-study: A multicenter diagnostic accuracy study of MRI in patients with suspected acute appendicitis. Study Protocol

    Directory of Open Access Journals (Sweden)

    Bossuyt Patrick MM

    2010-10-01

    the direct and indirect costs associated with treatment. Based on the high intrinsic contrast resolution of MRI, one might envision higher accuracy rates for MRI than for CT. If so, MRI could further decrease the number of unnecessary appendectomies and the number of missed appendicitis cases. Trial registration NTR2148

  6. Accuracy and Utility of Deformable Image Registration in {sup 68}Ga 4D PET/CT Assessment of Pulmonary Perfusion Changes During and After Lung Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Hardcastle, Nicholas, E-mail: nick.hardcastle@gmail.com [Department of Physical Sciences, Peter MacCallum Cancer Centre, East Melbourne (Australia); Centre for Medical Radiation Physics, University of Wollongong, Wollongong (Australia); Hofman, Michael S. [Molecular Imaging, Centre for Cancer Imaging, Peter MacCallum Cancer Centre, Melbourne (Australia); Hicks, Rodney J. [Molecular Imaging, Centre for Cancer Imaging, Peter MacCallum Cancer Centre, Melbourne (Australia); Department of Medicine, University of Melbourne, Melbourne (Australia); Callahan, Jason [Molecular Imaging, Centre for Cancer Imaging, Peter MacCallum Cancer Centre, Melbourne (Australia); Kron, Tomas [Department of Medical Imaging and Radiation Sciences, Monash University, Clayton (Australia); The Sir Peter MacCallum Department of Oncology, Melbourne University, Victoria (Australia); MacManus, Michael P.; Ball, David L. [Division of Radiation Oncology and Cancer Imaging, Peter MacCallum Cancer Centre, East Melbourne (Australia); The Sir Peter MacCallum Department of Oncology, University of Melbourne, Melbourne (Australia); Jackson, Price [Department of Physical Sciences, Peter MacCallum Cancer Centre, East Melbourne (Australia); Siva, Shankar [Division of Radiation Oncology and Cancer Imaging, Peter MacCallum Cancer Centre, East Melbourne (Australia)

    2015-09-01

    Purpose: Measuring changes in lung perfusion resulting from radiation therapy dose requires registration of the functional imaging to the radiation therapy treatment planning scan. This study investigates registration accuracy and utility for positron emission tomography (PET)/computed tomography (CT) perfusion imaging in radiation therapy for non–small cell lung cancer. Methods: {sup 68}Ga 4-dimensional PET/CT ventilation-perfusion imaging was performed before, during, and after radiation therapy for 5 patients. Rigid registration and deformable image registration (DIR) using B-splines and Demons algorithms was performed with the CT data to obtain a deformation map between the functional images and planning CT. Contour propagation accuracy and correspondence of anatomic features were used to assess registration accuracy. Wilcoxon signed-rank test was used to determine statistical significance. Changes in lung perfusion resulting from radiation therapy dose were calculated for each registration method for each patient and averaged over all patients. Results: With B-splines/Demons DIR, median distance to agreement between lung contours reduced modestly by 0.9/1.1 mm, 1.3/1.6 mm, and 1.3/1.6 mm for pretreatment, midtreatment, and posttreatment (P<.01 for all), and median Dice score between lung contours improved by 0.04/0.04, 0.05/0.05, and 0.05/0.05 for pretreatment, midtreatment, and posttreatment (P<.001 for all). Distance between anatomic features reduced with DIR by median 2.5 mm and 2.8 for pretreatment and midtreatment time points, respectively (P=.001) and 1.4 mm for posttreatment (P>.2). Poorer posttreatment results were likely caused by posttreatment pneumonitis and tumor regression. Up to 80% standardized uptake value loss in perfusion scans was observed. There was limited change in the loss in lung perfusion between registration methods; however, Demons resulted in larger interpatient variation compared with rigid and B-splines registration

  7. Accuracy of spiral CT and 3D reconstruction in the detection of acute pulmonary embolism - development of an animal model using porcine lungs and technical specimens. Development of an animal model using porcine lungs and technical specimens

    International Nuclear Information System (INIS)

    Ries, B.G.; Kauczor, H.U.; Thelen, M.; Konerding, M.A.

    2001-01-01

    Purpose: To develop a model for simulation the CT morphologic situation of acute pulmonary embolism, to evaluate the accuracy of spiral CT and 3D reconstruction in the detection of artificial emboli and to investigate the influence of the orientation of emboli depending on z-axis orientation. Materials and Methods: Standardized artificial emboli made of wax and of defined size and shape were positioned into the pulmonary arteries of porcine lungs. Castings of the embolized pulmonary arterial trees were made by injection of a special opaque resin. After performance of spiral CT the data sets of the emboli and the pulmonary arteries were post-processed. The 3D segmentations were compared with the anatomic preparation to evaluate the accuracy of spiral CT/3D reconstruction-technique. Technical specimens simulating CT-morphology of acute embolized vessels underwent spiral CT in six different positions with respect to the z-axis. The CT data were reconstructed using a standardized and a contrastadapted method with interactive correction. The 3D emboli were analysed under qualitative aspects, and measurements of their extent were done. Results: In nearly 91%, there was complete agreement between CT and the corresponding findings at the anatomical preparation. Measurements of the 3D reconstructed technical specimens showed discrepancies of shape and size in dependence of the size of the original preparation, orientation and reconstruction technique. Overestimation up to 4 mm and underestimation to 2,2 mm were observed. Measurements of preparations with heights from 14 to 26 mm showed variances of ±1,5 mm (∝6-11%). Conclusion: The presented models are suitable to simulate CT morphology of acute pulmonary embolism under ex-vivo conditions. Accuracy in the detection of artificial emboli using spiral CT/3D reconstruction is affected by localization, size and orientation of the emboli and the reconstruction technique. (orig.) [de

  8. Influence of density and mean atomic number on CT attenuation corrected PET: Phantom studies

    International Nuclear Information System (INIS)

    Maintas, D.; Houzard, C.; Galy, G.; Maintas, C.; Itti, R.; Cachin, F.; Mognetti, Th.; Slosman

    2007-01-01

    Aim: the aim of this work is to study the influence of medium density on the CT or external source attenuation corrected images, by simulation on a phantom, with various positron emission tomographs. Material and method: a series of experiments on a cylindrical phantom filled with water marked with [18 F]-FDG, containing six vials filled per pair with mediums of different densities or solutions of KI, CaCl 2 and saccharose with various densities, was carried out under comparable conditions on three different tomographs. In only one of the vials of each pair, an identical radioactivity of [18 F]-FDG was added, three to five fold the surrounding activity. The reconstructions and attenuation corrections suggested by the manufacturers, were carried out under the usual conditions of each site. The activity of each structure was estimated by the methods of profiles and regions of interest, on the non attenuation corrected images (N.A.C.), the images corrected by CT (C.T.A.C.), and/or external source (G.P.A.C.). Results: with all three tomographs, the activities estimated on the N.A.C. images present an inverse correlation to the medium density (important absorption by dense material). On C.T.A.C. images, we observed with only two of the three tomographs, an overestimation of the activity in the 'radioactive' vials, depending on the medium mean Z number and density (over correction), and a artifactual 'activity' in the denser 'cold' vial (incorrect attenuation correction. The dense saccharose solutions, with non elevated Z number, do not affect the CT attenuation correction. (authors)

  9. STACK NUMBER INFLUENCE ON THE ACCURACY OF ASTER GDEM (V2

    Directory of Open Access Journals (Sweden)

    S. M. J. Mirzadeh

    2017-09-01

    Full Text Available In this research, the influence of stack number (STKN on the accuracy of Advanced Spaceborne Thermal Emission and Reflection Radiometer (ASTER Global DEM (GDEM has been investigated. For this purpose, two data sets of ASTER and Reference DEMs from two study areas with various topography (Bomehen and Tazehabad were used. The Results show that in both study areas, STKN of 19 results in minimum error so that this minimum error has small difference with other STKN. The analysis of slope, STKN, and error values shows that there is no strong correlation between these parameters in both study areas. For example, the value of mean absolute error increase by changing the topography and the increase of slope values and height on cells but, the changes in STKN has no important effect on error values. Furthermore, according to high values of STKN, effect of slope on elevation accuracy has practically decreased. Also, there is no great correlation between the residual and STKN in ASTER GDEM.

  10. Diagnosis of gastric cancers by CT

    International Nuclear Information System (INIS)

    Zhu Jianbing; Gong Jianping; Huan Jian

    1999-01-01

    Forty two cases of gastric cancers were reviewed. The cancer had been examined by CT and was confirmed by operation and pathology. The diagnostic results of gastric cancers obtained by CT were compared with that from GI and fibro-gastroscopy examination. The results showed that the preparation of gastrointestinal tract before CT examination was important in the CT diagnosis of gastric cancer. CT in diagnosis of focus of gastric cancer and organ invasion is better than Gl and Fibro-gastroscopy and accuracy in diagnosis of gastric cancers is near to that of GI examination

  11. Diagnosis of Grave's disease with pulmonary hypertension on chest CT.

    Science.gov (United States)

    Lee, Hwa Yeon; Yoo, Seung Min; Kim, Hye Rin; Chun, Eun Ju; White, Charles S

    To evaluate the diagnostic accuracy of chest CT findings to diagnose Grave's disease in pulmonary hypertension. We retrospectively evaluated chest CT and the medical records of 13 patients with Grave's disease with (n=6) or without pulmonary hypertension (n=7) and in 17 control patients. Presence of iso-attenuation of diffusely enlarged thyroid glands compared with adjacent neck muscle on non-enhanced CT as a diagnostic clue of Grave's disease, and assessment of pulmonary hypertension on CT has high diagnostic accuracy. Chest CT has the potential to diagnose Grave's disease with pulmonary hypertension in the absence of other information. Copyright © 2017 Elsevier Inc. All rights reserved.

  12. Development and validation of an automated and marker-free CT-based spatial analysis method (CTSA) for assessment of femoral hip implant migration: In vitro accuracy and precision comparable to that of radiostereometric analysis (RSA).

    Science.gov (United States)

    Scheerlinck, Thierry; Polfliet, Mathias; Deklerck, Rudi; Van Gompel, Gert; Buls, Nico; Vandemeulebroucke, Jef

    2016-01-01

    We developed a marker-free automated CT-based spatial analysis (CTSA) method to detect stem-bone migration in consecutive CT datasets and assessed the accuracy and precision in vitro. Our aim was to demonstrate that in vitro accuracy and precision of CTSA is comparable to that of radiostereometric analysis (RSA). Stem and bone were segmented in 2 CT datasets and both were registered pairwise. The resulting rigid transformations were compared and transferred to an anatomically sound coordinate system, taking the stem as reference. This resulted in 3 translation parameters and 3 rotation parameters describing the relative amount of stem-bone displacement, and it allowed calculation of the point of maximal stem migration. Accuracy was evaluated in 39 comparisons by imposing known stem migration on a stem-bone model. Precision was estimated in 20 comparisons based on a zero-migration model, and in 5 patients without stem loosening. Limits of the 95% tolerance intervals (TIs) for accuracy did not exceed 0.28 mm for translations and 0.20° for rotations (largest standard deviation of the signed error (SD(SE)): 0.081 mm and 0.057°). In vitro, limits of the 95% TI for precision in a clinically relevant setting (8 comparisons) were below 0.09 mm and 0.14° (largest SD(SE): 0.012 mm and 0.020°). In patients, the precision was lower, but acceptable, and dependent on CT scan resolution. CTSA allows detection of stem-bone migration with an accuracy and precision comparable to that of RSA. It could be valuable for evaluation of subtle stem loosening in clinical practice.

  13. The use of CT scan in the pre-operative staging of bronchogenic carcinoma

    International Nuclear Information System (INIS)

    Pada, C.C.

    1992-01-01

    Surgery remains the treatment of choice in patients with localized bronchogenic carcinoma. Pre-operative identification of inoperability spares the patient from unnecessary surgery. This prospective study was carried out to determine the correctness of judgement regarding a patient's operability or inoperability based on the pre-operative staging of CT scan; to find out the sensitivity, specificity and overall accuracy of the CT scan in estimating tumor description, nodal status and metastatic spread to the chest. Staging was done by 3 senior radiologists aware of the diagnosis. Both the surgical and histopathologic findings and staging were gathered and used as measurement of truth in arriving at the CT scan's accuracy. Overall accuracy rate of CT scan in determining operability or inoperability is 80%; tumor description accuracy of assessment is 87% and nodal status estimation has an accuracy of 60%. Sensitivity of CT scan is assessment of metastatic spread to the chest is 93%. There is no statistically significant difference in the judgement of operability or interpretability by CT scan compared to surgical and histopathologic results. The CT scan is recommended as a valuable tool in the pre-operative staging of patients with bronchogenic carcinoma who are candidates for surgery. (auth.). 21 refs.; 8 tabs

  14. Comparative analysis of multi-slice spiral CT and positron emission tomography-CT in evaluation of axillary lymph nodes in breast cancer patients

    International Nuclear Information System (INIS)

    Sun Xianchang; Zhang Ruyi; Liu Qingwei; Zhao Suhong; Zu Degui; Li Xin

    2008-01-01

    Objective: To evaluate and compare spiral CT and positron emission tomography-CT (PET-CT) in characterization of of axillary lymph nodes in breast cancer patients. Methods: Forty patients with pathologically proven breast cancer underwent contrast-enhanced spiral CT of the breast and axilla, 13 of them also underwent PET-CT examination. One hundred and fifty-eight axillary lymph nodes were found in the 40 patients through contrast enhanced spiral CT, while 57 lymph nodes were found in the 13 patients through PET-CT. Three radiologists rated the lymph nodes found in CT images on a five-point scale. If the score was equal to or greater than 3, it was defined as positive (metastatic), otherwise negative. Visual observation and semiquantitative analysis were used to classify lymph nodes in PET-CT images. The results of spiral CT observation and PET-CT observation of lymph nodes were compared with pathological results. The relative value of CT and PET-CT was analyzed. Exact probability statistics were employed. Results: One hundred and fifty eight lymph nodes of 40 patients were detected by spiral CT, 91 of them were diagnosed as positive and 67 as negative Among the lymph nodes found in spiral CT, 99 were positive and 59 were negative pathologicall. A total of 57 lymph nodes were found by PET-CT. Thirty-nine of them were defined as positive and 18 as negative. Among the lymph nodes found in PET-CT, 39 were positive and 18 were negative pathologically. The sensitivity, specificity, accuracy, positive and negative predictive values in CT prediction in axillary lymph nodes metastases were 88.89%, 94.91%, 91.14%, 96.70%, and 83.58%, respectively. The sensitivity, specificity, accuracy, positive and negative predictive values in PET-CT prediction in axillary lymph nodes metastases were 97.44%, 94.44%, 96.49%, 97.44%, and 94.44%, respectively. PET-CT had no significant difference with spiral CT in sensitivity, accuracy, positive and negative predictive values for detection

  15. Treatment planning using MRI data: an analysis of the dose calculation accuracy for different treatment regions

    Directory of Open Access Journals (Sweden)

    Karlsson Mikael

    2010-06-01

    Full Text Available Abstract Background Because of superior soft tissue contrast, the use of magnetic resonance imaging (MRI as a complement to computed tomography (CT in the target definition procedure for radiotherapy is increasing. To keep the workflow simple and cost effective and to reduce patient dose, it is natural to strive for a treatment planning procedure based entirely on MRI. In the present study, we investigate the dose calculation accuracy for different treatment regions when using bulk density assignments on MRI data and compare it to treatment planning that uses CT data. Methods MR and CT data were collected retrospectively for 40 patients with prostate, lung, head and neck, or brain cancers. Comparisons were made between calculations on CT data with and without inhomogeneity corrections and on MRI or CT data with bulk density assignments. The bulk densities were assigned using manual segmentation of tissue, bone, lung, and air cavities. Results The deviations between calculations on CT data with inhomogeneity correction and on bulk density assigned MR data were small. The maximum difference in the number of monitor units required to reach the prescribed dose was 1.6%. This result also includes effects of possible geometrical distortions. Conclusions The dose calculation accuracy at the investigated treatment sites is not significantly compromised when using MRI data when adequate bulk density assignments are made. With respect to treatment planning, MRI can replace CT in all steps of the treatment workflow, reducing the radiation exposure to the patient, removing any systematic registration errors that may occur when combining MR and CT, and decreasing time and cost for the extra CT investigation.

  16. Diagnostic accuracy of dual energy CT angiography in patients with diabetes mellitus; Diagnostische Genauigkeit der Dual-energy-CT-Angiographie bei Patienten mit Diabetes mellitus

    Energy Technology Data Exchange (ETDEWEB)

    Schabel, C.; Bongers, M.N.; Syha, R. [Klinikum der Eberhard-Karls-Universitaet, Abteilung fuer Diagnostische und Interventionelle Radiologie, Tuebingen (Germany); Klinikum der Eberhard-Karls-Universitaet, Sektion fuer Experimentelle Radiologie der Abteilung fuer Diagnostische und Interventionelle Radiologie, Tuebingen (Germany); Ketelsen, D.; Homann, G.; Notohamiprodjo, M.; Nikolaou, K.; Bamberg, F. [Klinikum der Eberhard-Karls-Universitaet, Abteilung fuer Diagnostische und Interventionelle Radiologie, Tuebingen (Germany); Thomas, C. [Universitaetsklinikum Duesseldorf, Abteilung fuer Diagnostische und Interventionelle Radiologie, Duesseldorf (Germany)

    2015-04-01

    Peripheral arterial disease (PAD) represents a major and highly prevalent complication in patients with diabetes mellitus. The diagnostic, non-invasive work-up by computed tomography angiography (CTA) is limited in the presence of extensive calcification. The aim of the study was to determine the diagnostic accuracy of dual energy CTA (DE-CTA) for the detection and characterization of PAD in patients with diabetes mellitus. In this study 30 diabetic patients with suspected or known PAD were retrospectively included in the analysis. All subjects underwent DE-CTA (Somatom Definition Flash, Siemens Healthcare, Erlangen, Germany) prior to invasive angiography, which served as the reference standard. Blinded analysis included assessment of the presence and degree of peripheral stenosis on curved multiplanar reformatting (MPR) and maximum intensity projections (MIP). Conventional measures of diagnostic accuracy were derived. Among the 30 subjects included in the analysis (83 % male, mean age 70.0 ± 10.5 years, 83 % diabetes type 2), the prevalence of critical stenosis in 331 evaluated vessel segments was high (30 %). Dual energy CT identified critical stenoses with a high sensitivity and good specificity using curved MPR (100 % and 93.1 %, respectively) and MIP images (99 % and 91.8 %, respectively). In stratified analysis, the diagnostic accuracy was higher for stenosis pertaining to the pelvic and thigh vessels as compared with the lower extremities (curved MPR accuracy 97.1 % vs. 99.2 vs. 90.9 %; respectively, p < 0.001). The use of DE-CTA allows reliable detection and characterization of peripheral arterial stenosis in patients with diabetes mellitus with higher accuracy in vessels in the pelvic and thigh regions compared with the vessels in the lower legs. (orig.) [German] Die periphere arterielle Verschlusskrankheit (PAVK) ist eine wesentliche Komplikation des Diabetes mellitus und stellt aufgrund ausgepraegter Gefaessverkalkungen eine diagnostische

  17. An evaluation of three commercially available metal artifact reduction methods for CT imaging

    International Nuclear Information System (INIS)

    Huang, Jessie Y; Kerns, James R; Balter, Peter A; Followill, David S; Mirkovic, Dragan; Howell, Rebecca M; Kry, Stephen F; Nute, Jessica L; Liu, Xinming; Stingo, Francesco C

    2015-01-01

    Three commercial metal artifact reduction methods were evaluated for use in computed tomography (CT) imaging in the presence of clinically realistic metal implants: Philips O-MAR, GE’s monochromatic gemstone spectral imaging (GSI) using dual-energy CT, and GSI monochromatic imaging with metal artifact reduction software applied (MARs). Each method was evaluated according to CT number accuracy, metal size accuracy, and streak artifact severity reduction by using several phantoms, including three anthropomorphic phantoms containing metal implants (hip prosthesis, dental fillings and spinal fixation rods). All three methods showed varying degrees of success for the hip prosthesis and spinal fixation rod cases, while none were particularly beneficial for dental artifacts. Limitations of the methods were also observed. MARs underestimated the size of metal implants and introduced new artifacts in imaging planes beyond the metal implant when applied to dental artifacts, and both the O-MAR and MARs algorithms induced artifacts for spinal fixation rods in a thoracic phantom. Our findings suggest that all three artifact mitigation methods may benefit patients with metal implants, though they should be used with caution in certain scenarios. (paper)

  18. Precision and accuracy in CT attenuation measurement of vascular wall using region-of-interest supported by differentiation curve

    International Nuclear Information System (INIS)

    Suzuki, Shigeru; Kidouchi, Takashi; Kuwahara, Sadatoshi; Vembar, Mani; Takei, Ryoji; Yamamoto, Asako

    2012-01-01

    Objectives: To evaluate the precision and accuracy in CT attenuation measurement of vascular wall using region-of-interest (ROI) supported by differentiation curves. Study design: We used vascular models (actual attenuation value of the wall: 87 HU) with wall thicknesses of 1.5, 1.0, or 0.5 mm, filled with contrast material of 250, 348, or 436 HU. The nine vascular models were scanned with a 64-detector CT. The wall attenuation values were measured using three sizes (diameter: 0.5, 1.0, and 1.5 mm) of ROIs without differentiation curves. Sixteen measurements were repeated for each vascular model by each of two operators. Measurements supported by differentiation curves were also performed. We used analyses of variance with repeated measures for the measured attenuations for each size of the ROI. Results: Without differentiation curves, there were significant differences in the attenuation values of the wall among the three densities of contrast material, and the attenuation values tended to be overestimated more as the contrast material density increased. Operator dependencies were also found in measurements for 0.5- and 1.5-mm thickness models. With differentiation curves, measurements were not possible for 0.5- and 1.0-mm thickness models. Using differentiation curves for 1.5-mm thickness models with a ROI of 1.0- or 1.5-mm diameter, the wall attenuations were not affected by the contrast material densities and were operator independent, measuring between 75 and 103 HU. Conclusions: The use of differentiation curves can improve the precision and accuracy in wall attenuation measurement using a ROI technique, while measurements for walls of ≤1.0 mm thickness are difficult.

  19. Evaluation of cervical lymph node metastasis in thyroid cancer patients using real-time CT navigated ultrasonography: preliminary study

    International Nuclear Information System (INIS)

    Na, Dae Kwon; Choi, Yoon Jung; Choi, Seon Hyeong; Kook, Shin Ho; Park, Hee Jin

    2015-01-01

    To evaluate the diagnostic accuracy of real-time neck computed tomography (CT)-guided ultrasonography (US) in detecting cervical neck lymph node metastasis (LNM) in patients with papillary thyroid cancer (PTC). We retrospectively reviewed data from 176 patients (mean age, 43 years; range, 23 to 74 years) with surgically confirmed PTC who underwent preoperative US, neck CT, and neck CTguided US. We then compared the sensitivities and diagnostic accuracies of each of the three above modalities in detecting cervical LNM. Preoperative US showed 17.3% sensitivity and 58.5% diagnostic accuracy in detecting central LNM compared with 64.3% sensitivity and 89.2% diagnostic accuracy in detecting lateral neck LNM. Neck CT showed 23.5% sensitivity and 55.7% diagnostic accuracy in detecting central LNM and 71.4% sensitivity with 90.9% diagnostic accuracy in detecting lateral neck LNM. CT-guided US exhibited 37.0% sensitivity and 63.1% diagnostic accuracy in detecting central LNM compared with 92.9% sensitivity and 96.0% diagnostic accuracy in detecting lateral LNM. CT-guided US showed higher diagnostic accuracy with superior sensitivity in detecting central and lateral LNM than did US (P<0.001, P=0.011) and CT (P=0.026, P=0.063). Neck CT-guided US is a more accurate technique with higher sensitivity for detecting cervical LNM than either US or CT alone. Therefore, our data indicate that neck CT-guided US is an especially useful technique in preoperative examinations.

  20. SU-F-207-09: Evaluating the Dosimetric Accuracy of Extended Field-Of-View CT Reconstructions Using Clinical Data with Real Patient Geometries

    Energy Technology Data Exchange (ETDEWEB)

    Shugard, E; Cheung, J; Pouliot, J; Chen, J [University of California San Francisco, San Francisco (United States); Mistry, N [Siemens Healthcare USA, Malvern, PA (United States)

    2015-06-15

    Purpose: To determine the accuracy of dose calculations performed with CT images reconstructed using extended field of view (FOV) algorithms. Methods: In this study we selected 6 radiotherapy patients (3 head and neck & 3 chest/pelvis) whose body circumferences extended past the 50 cm scan FOV in the treatment planning CT. Images acquired on a Siemens Sensation Open scanner, were reconstructed using the standard FOV (sFOV) and two different extended FOV algorithms, eFOV and HDFOV. A physician and dosimetrist identified the radiation target, critical organs, and external patient contour. A benchmark CT was created for each patient, consisting of an average of the 3 CT reconstructions with a density override applied to regions containing truncation artifacts, and was used to create an optimal radiation treatment plan. The plan was copied onto each reconstruction without density override and dose was recalculated. Results: The native sFOV dose calculations had the largest deviation from the benchmark (0.4 – 6.0%). Both the HDFOV and eFOV calculations showed improvement over the sFOV. For patients with a smooth patient contour, the HDFOV calculation had the least deviation from the benchmark (0.1–0.5%) compared to eFOV (0.4–1.8%). In cases with large amounts of tissue and irregular skin folds, the eFOV deviated the least from the benchmark (0.2–0.6%) compared to HDFOV (1.3–1.8%). It was observed that the eFOV scan has large image artifact in air with an average HU of −600 compared to the ideal of −1000. In these cases, the artifact in air appears to attenuate the dose and compensate for the missing tissue density. The HDFOV demonstrated minimal artifact in air. Conclusion: Both eFOV and HDFOV provide improved dose calculation accuracy. The HDFOV reconstruction provides a clearer patient border than the eFOV allowing for well-defined density overrides to be applied with minimal air artifact. This research was supported by Siemens Medical Solutions.

  1. Prospective versus retrospective ECG-gated multislice CT coronary angiography: A systematic review of radiation dose and diagnostic accuracy

    International Nuclear Information System (INIS)

    Sun Zhonghua; Ng, Kwan-Hoong

    2012-01-01

    Purpose: To perform a systematic review of the radiation dose and diagnostic accuracy of prospective versus retrospective ECG-gated multislice CT coronary angiography. Materials and methods: A search of Pubmed/Medline and Sciencedirect databases for English literature was performed to identify studies comparing prospective and retrospective ECG-gated multislice CT angiography in the diagnosis of coronary artery disease. Effective dose, dose length product, image quality and diagnostic value were compared between two groups of studies. Results: 22 studies were included for analysis. The mean effective dose of prospective ECG-gated scans was 4.5 mSv (95% CI: 3.6, 5.3 mSv), which is significantly lower than that of retrospective scans, which is 13.8 mSv (95% CI: 11.5, 16.0 mSv) (p < 0.001). The mean dose length product was 225 mGy cm (95% CI: 188, 262 mGy cm) and 822 mGy cm (95% CI: 630, 1013 mGy cm) for the prospective and retrospective ECG-gated scans, respectively, indicating a statistically significant difference between these two protocols (p < 0.0001). The mean sensitivity and specificity of multislice CT angiography in the diagnosis of coronary artery disease was 97.7% (95% CI: 93.7%, 100%) and 92.1% (95% CI: 87.2%, 97%) for prospective ECG-gated scans; 95.2% (95% CI: 91%, 99.5%) and 94.4% (95% CI: 88.5%, 100%) for retrospective ECG-gated scans, respectively, with no significant difference for sensitivity but significant difference for specificity (p = 0.047). Conclusion: Multislice CT coronary angiography with prospective ECG-gating leads to a significant reduction of radiation dose when compared to that of retrospective ECG-gating, while offering comparable image quality and diagnostic value.

  2. SU-E-J-219: Quantitative Evaluation of Motion Effects On Accuracy of Image-Guided Radiotherapy with Fiducial Markers Using CT Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Ali, I; Oyewale, S; Ahmad, S; Algan, O [University of Oklahoma Health Sciences, Oklahoma City, OK (United States); Alsbou, N [Department of Electrical and Computer Engineering, Ada, OH (United States)

    2014-06-01

    Purpose: To investigate quantitatively patient motion effects on the localization accuracy of image-guided radiation with fiducial markers using axial CT (ACT), helical CT (HCT) and cone-beam CT (CBCT) using modeling and experimental phantom studies. Methods: Markers with different lengths (2.5 mm, 5 mm, 10 mm, and 20 mm) were inserted in a mobile thorax phantom which was imaged using ACT, HCT and CBCT. The phantom moved with sinusoidal motion with amplitudes ranging 0–20 mm and a frequency of 15 cycles-per-minute. Three parameters that include: apparent marker lengths, center position and distance between the centers of the markers were measured in the different CT images of the mobile phantom. A motion mathematical model was derived to predict the variations in the previous three parameters and their dependence on the motion in the different imaging modalities. Results: In CBCT, the measured marker lengths increased linearly with increase in motion amplitude. For example, the apparent length of the 10 mm marker was about 20 mm when phantom moved with amplitude of 5 mm. Although the markers have elongated, the center position and the distance between markers remained at the same position for different motion amplitudes in CBCT. These parameters were not affected by motion frequency and phase in CBCT. In HCT and ACT, the measured marker length, center and distance between markers varied irregularly with motion parameters. The apparent lengths of the markers varied with inverse of the phantom velocity which depends on motion frequency and phase. Similarly the center position and distance between markers varied inversely with phantom speed. Conclusion: Motion may lead to variations in maker length, center position and distance between markers using CT imaging. These effects should be considered in patient setup using image-guided radiation therapy based on fiducial markers matching using 2D-radiographs or volumetric CT imaging.

  3. SU-E-J-219: Quantitative Evaluation of Motion Effects On Accuracy of Image-Guided Radiotherapy with Fiducial Markers Using CT Imaging

    International Nuclear Information System (INIS)

    Ali, I; Oyewale, S; Ahmad, S; Algan, O; Alsbou, N

    2014-01-01

    Purpose: To investigate quantitatively patient motion effects on the localization accuracy of image-guided radiation with fiducial markers using axial CT (ACT), helical CT (HCT) and cone-beam CT (CBCT) using modeling and experimental phantom studies. Methods: Markers with different lengths (2.5 mm, 5 mm, 10 mm, and 20 mm) were inserted in a mobile thorax phantom which was imaged using ACT, HCT and CBCT. The phantom moved with sinusoidal motion with amplitudes ranging 0–20 mm and a frequency of 15 cycles-per-minute. Three parameters that include: apparent marker lengths, center position and distance between the centers of the markers were measured in the different CT images of the mobile phantom. A motion mathematical model was derived to predict the variations in the previous three parameters and their dependence on the motion in the different imaging modalities. Results: In CBCT, the measured marker lengths increased linearly with increase in motion amplitude. For example, the apparent length of the 10 mm marker was about 20 mm when phantom moved with amplitude of 5 mm. Although the markers have elongated, the center position and the distance between markers remained at the same position for different motion amplitudes in CBCT. These parameters were not affected by motion frequency and phase in CBCT. In HCT and ACT, the measured marker length, center and distance between markers varied irregularly with motion parameters. The apparent lengths of the markers varied with inverse of the phantom velocity which depends on motion frequency and phase. Similarly the center position and distance between markers varied inversely with phantom speed. Conclusion: Motion may lead to variations in maker length, center position and distance between markers using CT imaging. These effects should be considered in patient setup using image-guided radiation therapy based on fiducial markers matching using 2D-radiographs or volumetric CT imaging

  4. The sensitivity of computed tomography (CT) scans in detecting trauma: are CT scans reliable enough for courtroom testimony?

    Science.gov (United States)

    Molina, D Kimberley; Nichols, Joanna J; Dimaio, Vincent J M

    2007-09-01

    Rapid and accurate recognition of traumatic injuries is extremely important in emergency room and surgical settings. Emergency departments depend on computed tomography (CT) scans to provide rapid, accurate injury assessment. We conducted an analysis of all traumatic deaths autopsied at the Bexar County Medical Examiner's Office in which perimortem medical imaging (CT scan) was performed to assess the reliability of the CT scan in detecting trauma with sufficient accuracy for courtroom testimony. Cases were included in the study if an autopsy was conducted, a CT scan was performed within 24 hours before death, and there was no surgical intervention. Analysis was performed to assess the correlation between the autopsy and CT scan results. Sensitivity, specificity, positive predictive value, and negative predictive value were defined for the CT scan based on the autopsy results. The sensitivity of the CT scan ranged from 0% for cerebral lacerations, cervical vertebral body fractures, cardiac injury, and hollow viscus injury to 75% for liver injury. This study reveals that CT scans are an inadequate detection tool for forensic pathologists, where a definitive diagnosis is required, because they have a low level of accuracy in detecting traumatic injuries. CT scans may be adequate for clinicians in the emergency room setting, but are inadequate for courtroom testimony. If the evidence of trauma is based solely on CT scan reports, there is a high possibility of erroneous accusations, indictments, and convictions.

  5. The CT (Hounsfield unit) number of brain tissue in healthy infants. A new reliable method for detection of possible degenerative disease.

    Science.gov (United States)

    Boris, P; Bundgaard, F; Olsen, A

    1987-01-01

    It is difficult to correlate CT Hounsfield unit (H. U.) numbers from one CT investigation to another and from one CT scanner to another, especially when dealing with small changes in the brain substance, as in degenerative brain diseases in children. By subtracting the mean value of cerebrospinal fluid (CSF) from the mean value of grey and white matter, it is possible to remove most of the errors due, for example, to maladjustments, short and long-term drift, X-ray fan, and detector asymmetry. Measurements of white and grey matter using these methods showed CT H. U. numbers changing from 15 H. U. to 22 H. U. in white matter and 23 H. U. to 30 H. U. in grey matter in 86 healthy infants aged 0-5 years. In all measurements, the difference between grey and white matter was exactly 8 H. U. The method has proven to be highly accurate and reproducible.

  6. Diagnostic accuracy of contrast-enhanced MR angiography and unenhanced proton MR imaging compared with CT pulmonary angiography in chronic thromboembolic pulmonary hypertension

    Energy Technology Data Exchange (ETDEWEB)

    Rajaram, Smitha [Unit of Academic Radiology, University of Sheffield, Sheffield (United Kingdom); Academic Unit of Radiology, C Floor, Royal Hallamshire Hospital, Sheffield (United Kingdom); Swift, Andrew J.; Wild, Jim M. [Unit of Academic Radiology, University of Sheffield, Sheffield (United Kingdom); Sheffield Cardiovascular Biomedical Research Unit, Sheffield (United Kingdom); Capener, David; Telfer, Adam [Unit of Academic Radiology, University of Sheffield, Sheffield (United Kingdom); Davies, Christine; Hill, Catherine [Sheffield Teaching Hospitals Trust, Department of Radiology, Sheffield (United Kingdom); Condliffe, Robin; Elliot, Charles; Kiely, David G. [Sheffield Pulmonary Vascular Disease Unit, Royal Hallamshire Hospital, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield (United Kingdom); Sheffield Cardiovascular Biomedical Research Unit, Sheffield (United Kingdom); Hurdman, Judith [Sheffield Pulmonary Vascular Disease Unit, Royal Hallamshire Hospital, Sheffield Teaching Hospitals NHS Foundation Trust, Sheffield (United Kingdom)

    2012-02-15

    To evaluate the diagnostic accuracy of contrast-enhanced MR angiography (CE-MRA) and the added benefit of unenhanced proton MR angiography compared with CT pulmonary angiography (CTPA) in patients with chronic thromboembolic disease (CTE). A 2 year retrospective study of 53 patients with chronic thromboembolic pulmonary hypertension who underwent CTPA and MRI for suspected pulmonary hypertension and a control group of 36 patients with no CT evidence of pulmonary embolism. The MRI was evaluated for CTE and the combined diagnostic accuracy of ce-MRA and unenhanced proton MRA was determined. CE-MRA generated lung perfusion maps were also assessed. The overall sensitivity and specificity of CE-MRA in diagnosing proximal and distal CTE were 98% and 94%, respectively. The sensitivity improved from 50% to 88% for central vessel disease when CE-MRA images were analysed with unenhanced proton MRA. The CE-MRA identified more stenoses (29/18), post-stenosis dilatation (23/7) and occlusions (37/29) compared with CTPA. The CE-MRA perfusion images showed a sensitivity of 92% for diagnosing CTE. CE-MRA has high sensitivity and specificity for diagnosing CTE. The sensitivity of CE-MRA for visualisation of adherent central and lobar thrombus significantly improves with the addition of unenhanced proton MRA which delineates the vessel wall. (orig.)

  7. Diagnostic accuracy of contrast-enhanced MR angiography and unenhanced proton MR imaging compared with CT pulmonary angiography in chronic thromboembolic pulmonary hypertension

    International Nuclear Information System (INIS)

    Rajaram, Smitha; Swift, Andrew J.; Wild, Jim M.; Capener, David; Telfer, Adam; Davies, Christine; Hill, Catherine; Condliffe, Robin; Elliot, Charles; Kiely, David G.; Hurdman, Judith

    2012-01-01

    To evaluate the diagnostic accuracy of contrast-enhanced MR angiography (CE-MRA) and the added benefit of unenhanced proton MR angiography compared with CT pulmonary angiography (CTPA) in patients with chronic thromboembolic disease (CTE). A 2 year retrospective study of 53 patients with chronic thromboembolic pulmonary hypertension who underwent CTPA and MRI for suspected pulmonary hypertension and a control group of 36 patients with no CT evidence of pulmonary embolism. The MRI was evaluated for CTE and the combined diagnostic accuracy of ce-MRA and unenhanced proton MRA was determined. CE-MRA generated lung perfusion maps were also assessed. The overall sensitivity and specificity of CE-MRA in diagnosing proximal and distal CTE were 98% and 94%, respectively. The sensitivity improved from 50% to 88% for central vessel disease when CE-MRA images were analysed with unenhanced proton MRA. The CE-MRA identified more stenoses (29/18), post-stenosis dilatation (23/7) and occlusions (37/29) compared with CTPA. The CE-MRA perfusion images showed a sensitivity of 92% for diagnosing CTE. CE-MRA has high sensitivity and specificity for diagnosing CTE. The sensitivity of CE-MRA for visualisation of adherent central and lobar thrombus significantly improves with the addition of unenhanced proton MRA which delineates the vessel wall. (orig.)

  8. CT interpretation of craniofacial anomalies: a comparative analysis by undergraduate dental students

    International Nuclear Information System (INIS)

    Gaia, Bruno Felipe; Perella, Andreia; Cara, Ana Claudia Ballet de; Antunes, Jose Leopoldo Ferreira; Cavalcanti, Marcelo Gusmao Paraiso

    2005-01-01

    The aim of this study was to evaluate the accuracy and reproducibility of computed tomography (CT) image interpretation made in axial slices (2D-CT) and 3D reconstructed images (3D-CT) of patients with craniofacial anomalies. The analyses were made by undergraduate dental students, and compared with the diagnoses considered upon surgical intervention. Computed tomography of 43 patients were analyzed independently by three calibrated examiners (undergraduate students) with, respectively, one, two, and three semesters of experience in craniofacial CT training and interpretation. The analysis of 2D-CT and 3D-CT images were performed at distinct times using an independent workstation associated with a specific computer graphics software for volumetric images. The analysis of inter-examiner agreement and of the agreement between observers and the gold standard was performed using the Kappa test. The accuracy evaluation presented a progressively higher value for examiners with progressively broader experience in 2D-CT and 3D-CT image interpretation. 3D-CT analyses allowed a higher inter-examiner agreement (1 - 0.896) than 2D-CT analyses (1 - 0.614). 3D-CT was considered more precise and accurate than 2D-CT for all students' evaluations. The reproducibility and accuracy varied according to the experience in CT interpretation, and the most experienced student achieved results closer to the gold standard. (author)

  9. PET/CT-guided percutaneous liver mass biopsies and ablations: Targeting accuracy of a single 20 s breath-hold PET acquisition

    International Nuclear Information System (INIS)

    Shyn, P.B.; Tatli, S.; Sahni, V.A.; Sadow, C.A.; Forgione, K.; Mauri, G.; Morrison, P.R.; Catalano, P.J.; Silverman, S.G.

    2014-01-01

    Aim: To determine whether a single 20 s breath-hold positron-emission tomography (PET) acquisition obtained during combined PET/computed tomography (CT)-guided percutaneous liver biopsy or ablation procedures has the potential to target 2-[ 18 F]-fluoro-2-deoxy-D-glucose (FDG)-avid liver masses as accurately as up to 180 s breath-hold PET acquisitions. Materials and methods: This retrospective study included 10 adult patients with 13 liver masses who underwent FDG PET/CT-guided percutaneous biopsies (n = 5) or ablations (n = 5). PET was acquired as nine sequential 20 s, monitored, same-level breath-hold frames and CT was acquired in one monitored breath-hold. Twenty, 40, 60, and 180 s PET datasets were reconstructed. Two blinded readers marked tumour centres on randomized PET and CT datasets. Three-dimensional spatial localization differences between PET datasets and either 180 s PET or CT were analysed using multiple regression analyses. Statistical tests were two-sided and p < 0.05 was considered significant. Results: Targeting differences between 20 s PET and 180 s PET ranged from 0.7–20.3 mm (mean 5.3 ± 4.4 mm; median 4.3) and were not statistically different from 40 or 60 s PET (p = 0.74 and 0.91, respectively). Targeting differences between 20 s PET and CT ranged from 1.4–36 mm (mean 9.6 ± 7.1 mm; median 8.2 mm) and were not statistically different from 40, 60, or 180 s PET (p = 0.84, 0.77, and 0.35, respectively). Conclusion: Single 20 s breath-hold PET acquisitions from PET/CT-guided percutaneous liver procedures have the potential to target FDG-avid liver masses with equivalent accuracy to 180 s summed, breath-hold PET acquisitions and may facilitate strategies that improve image registration and shorten procedure times

  10. Utility of CT-guided abdominal aspiration procedures

    International Nuclear Information System (INIS)

    Sundaram, M.; Wolverson, M.K.; Heiberg, E.; Pilla, T.; Vas, W.G.; Shields, J.B.

    1982-01-01

    Over 200 consecutive diagnostic needle aspiration procedures of the abdomen were performed under computed tomographic (CT) guidance. Biopsies were done of the liver in 88 patients, the pancreas in 28, the kidney in 20, and the retroperitoneum in 32; 30 underwent an aspiration procedure for characterization of an intraabdominal fluid collection. Accuracy of diagnosis was very high for hepatic (99%) and renal (100%) biopsies and in characterization of fluid collections (100%). Accuracy for retroperitoneal biopsy was 87.5% and for pancreatic biopsy 82%. Overall accuracy for all sites was 95%. There were one false-negative diagnosis for the liver and five false-negative diagnoses for the pancreas. Insufficient material was obtained for diagnosis in four instances of retroperitoneal biopsy. There were no false positives. The technique is facilitated by rapid CT scan time, large aperture gantry, and rapid CT image reconstruction. Twenty gauge needles were used more frequently than 22 gauge needles because of their greater rigidity and ease of control. CT-guided diagnostic aspiration procedures are particularly useful for diagnosis of small, deep-seated lesions and in evaluation of lesions found in severely ill patients. An experienced cytologist is essential to the success of the technique

  11. Influence of metallic dental implants and metal artefacts on dose calculation accuracy.

    Science.gov (United States)

    Maerz, Manuel; Koelbl, Oliver; Dobler, Barbara

    2015-03-01

    Metallic dental implants cause severe streaking artefacts in computed tomography (CT) data, which inhibit the correct representation of shape and density of the metal and the surrounding tissue. The aim of this study was to investigate the impact of dental implants on the accuracy of dose calculations in radiation therapy planning and the benefit of metal artefact reduction (MAR). A second aim was to determine the treatment technique which is less sensitive to the presence of metallic implants in terms of dose calculation accuracy. Phantoms consisting of homogeneous water equivalent material surrounding dental implants were designed. Artefact-containing CT data were corrected using the correct density information. Intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) plans were calculated on corrected and uncorrected CT data and compared to 2-dimensional dose measurements using GafChromic™ EBT2 films. For all plans the accuracy of dose calculations is significantly higher if performed on corrected CT data (p = 0.015). The agreement of calculated and measured dose distributions is significantly higher for VMAT than for IMRT plans for calculations on uncorrected CT data (p = 0.011) as well as on corrected CT data (p = 0.029). For IMRT and VMAT the application of metal artefact reduction significantly increases the agreement of dose calculations with film measurements. VMAT was found to provide the highest accuracy on corrected as well as on uncorrected CT data. VMAT is therefore preferable over IMRT for patients with metallic implants, if plan quality is comparable for the two techniques.

  12. Influence of metallic dental implants and metal artefacts on dose calculation accuracy

    International Nuclear Information System (INIS)

    Maerz, Manuel; Koelbl, Oliver; Dobler, Barbara

    2015-01-01

    Metallic dental implants cause severe streaking artefacts in computed tomography (CT) data, which inhibit the correct representation of shape and density of the metal and the surrounding tissue. The aim of this study was to investigate the impact of dental implants on the accuracy of dose calculations in radiation therapy planning and the benefit of metal artefact reduction (MAR). A second aim was to determine the treatment technique which is less sensitive to the presence of metallic implants in terms of dose calculation accuracy. Phantoms consisting of homogeneous water equivalent material surrounding dental implants were designed. Artefact-containing CT data were corrected using the correct density information. Intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) plans were calculated on corrected and uncorrected CT data and compared to 2-dimensional dose measurements using GafChromic trademark EBT2 films. For all plans the accuracy of dose calculations is significantly higher if performed on corrected CT data (p = 0.015). The agreement of calculated and measured dose distributions is significantly higher for VMAT than for IMRT plans for calculations on uncorrected CT data (p = 0.011) as well as on corrected CT data (p = 0.029). For IMRT and VMAT the application of metal artefact reduction significantly increases the agreement of dose calculations with film measurements. VMAT was found to provide the highest accuracy on corrected as well as on uncorrected CT data. VMAT is therefore preferable over IMRT for patients with metallic implants, if plan quality is comparable for the two techniques. (orig.) [de

  13. The accuracy of three-dimensional fused deposition modeling (FDM) compared with three-dimensional CT-Scans on the measurement of the mandibular ramus vertical length, gonion-menton length, and gonial angle

    Science.gov (United States)

    Savitri, I. T.; Badri, C.; Sulistyani, L. D.

    2017-08-01

    Presurgical treatment planning plays an important role in the reconstruction and correction of defects in the craniomaxillofacial region. The advance of solid freeform fabrication techniques has significantly improved the process of preparing a biomodel using computer-aided design and data from medical imaging. Many factors are implicated in the accuracy of the 3D model. To determine the accuracy of three-dimensional fused deposition modeling (FDM) models compared with three-dimensional CT scans in the measurement of the mandibular ramus vertical length, gonion-menton length, and gonial angle. Eight 3D models were produced from the CT scan data (DICOM file) of eight patients at the Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, University of Indonesia, Cipto Mangunkusumo Hospital. Three measurements were done three times by two examiners. The measurements of the 3D CT scans were made using OsiriX software, while the measurements of the 3D models were made using a digital caliper and goniometry. The measurement results were then compared. There is no significant difference between the measurements of the mandibular ramus vertical length, gonion-menton length, and gonial angle using 3D CT scans and FDM 3D models. FDM 3D models are considered accurate and are acceptable for clinical applications in dental and craniomaxillofacial surgery.

  14. The status of appendiceal CT in our hospital

    International Nuclear Information System (INIS)

    Honda, Yukiko; Nakanishi, Tadashi; Tagawa, Kohei; Akagi, Shinji; Kurisu, Yoshihiro; Ito, Katsuhide

    2006-01-01

    Images of 188 cases (112 males and 76 females; 5-88 years old, average 36) of patients with suspected appendicitis were retrospectively examined in the document (October 2003-July 2005). CT was conducted by 16-multidetector-row CT apparatus (MDCT) (Somatom Sensation Cardiac, Siemens) and evaluated the disease as positive, equivocal or negative. The number of positive diagnosis was 63, equivocal 5, and negative 120. Forty-four positive patients received the operation, 42 of whom were found to have appendicitis. Two out of 5 equivocal patients underwent the surgery and appendicitis was diagnosed in both. Ten negative patients received the operation through re-examination of their images, 6 of whom were found to have appendicitis. As above MDCT has improved the accuracy of diagnosis of acute appendicitis, indicating its usefulness in this field. (T.I.)

  15. Estimating local noise power spectrum from a few FBP-reconstructed CT scans

    Energy Technology Data Exchange (ETDEWEB)

    Zeng, Rongping, E-mail: rongping.zeng@fda.hhs.gov; Gavrielides, Marios A.; Petrick, Nicholas; Sahiner, Berkman; Li, Qin; Myers, Kyle J. [Division of Imaging, Diagnostics, and Software Reliability, Office of Science and Engineering Laboratories, CDRH, FDA, Silver Spring, Maryland 20993 (United States)

    2016-01-15

    Purpose: Traditional ways to estimate 2D CT noise power spectrum (NPS) involve an ensemble average of the power spectrums of many noisy scans. When only a few scans are available, regions of interest are often extracted from different locations to obtain sufficient samples to estimate the NPS. Using image samples from different locations ignores the nonstationarity of CT noise and thus cannot accurately characterize its local properties. The purpose of this work is to develop a method to estimate local NPS using only a few fan-beam CT scans. Methods: As a result of FBP reconstruction, the CT NPS has the same radial profile shape for all projection angles, with the magnitude varying with the noise level in the raw data measurement. This allows a 2D CT NPS to be factored into products of a 1D angular and a 1D radial function in polar coordinates. The polar separability of CT NPS greatly reduces the data requirement for estimating the NPS. The authors use this property and derive a radial NPS estimation method: in brief, the radial profile shape is estimated from a traditional NPS based on image samples extracted at multiple locations. The amplitudes are estimated by fitting the traditional local NPS to the estimated radial profile shape. The estimated radial profile shape and amplitudes are then combined to form a final estimate of the local NPS. We evaluate the accuracy of the radial NPS method and compared it to traditional NPS methods in terms of normalized mean squared error (NMSE) and signal detectability index. Results: For both simulated and real CT data sets, the local NPS estimated with no more than six scans using the radial NPS method was very close to the reference NPS, according to the metrics of NMSE and detectability index. Even with only two scans, the radial NPS method was able to achieve a fairly good accuracy. Compared to those estimated using traditional NPS methods, the accuracy improvement was substantial when a few scans were available

  16. Spectral CT metal artifact reduction with an optimization-based reconstruction algorithm

    Science.gov (United States)

    Gilat Schmidt, Taly; Barber, Rina F.; Sidky, Emil Y.

    2017-03-01

    Metal objects cause artifacts in computed tomography (CT) images. This work investigated the feasibility of a spectral CT method to reduce metal artifacts. Spectral CT acquisition combined with optimization-based reconstruction is proposed to reduce artifacts by modeling the physical effects that cause metal artifacts and by providing the flexibility to selectively remove corrupted spectral measurements in the spectral-sinogram space. The proposed Constrained `One-Step' Spectral CT Image Reconstruction (cOSSCIR) algorithm directly estimates the basis material maps while enforcing convex constraints. The incorporation of constraints on the reconstructed basis material maps is expected to mitigate undersampling effects that occur when corrupted data is excluded from reconstruction. The feasibility of the cOSSCIR algorithm to reduce metal artifacts was investigated through simulations of a pelvis phantom. The cOSSCIR algorithm was investigated with and without the use of a third basis material representing metal. The effects of excluding data corrupted by metal were also investigated. The results demonstrated that the proposed cOSSCIR algorithm reduced metal artifacts and improved CT number accuracy. For example, CT number error in a bright shading artifact region was reduced from 403 HU in the reference filtered backprojection reconstruction to 33 HU using the proposed algorithm in simulation. In the dark shading regions, the error was reduced from 1141 HU to 25 HU. Of the investigated approaches, decomposing the data into three basis material maps and excluding the corrupted data demonstrated the greatest reduction in metal artifacts.

  17. 4D-CT Lung registration using anatomy-based multi-level multi-resolution optical flow analysis and thin-plate splines.

    Science.gov (United States)

    Min, Yugang; Neylon, John; Shah, Amish; Meeks, Sanford; Lee, Percy; Kupelian, Patrick; Santhanam, Anand P

    2014-09-01

    The accuracy of 4D-CT registration is limited by inconsistent Hounsfield unit (HU) values in the 4D-CT data from one respiratory phase to another and lower image contrast for lung substructures. This paper presents an optical flow and thin-plate spline (TPS)-based 4D-CT registration method to account for these limitations. The use of unified HU values on multiple anatomy levels (e.g., the lung contour, blood vessels, and parenchyma) accounts for registration errors by inconsistent landmark HU value. While 3D multi-resolution optical flow analysis registers each anatomical level, TPS is employed for propagating the results from one anatomical level to another ultimately leading to the 4D-CT registration. 4D-CT registration was validated using target registration error (TRE), inverse consistency error (ICE) metrics, and a statistical image comparison using Gamma criteria of 1 % intensity difference in 2 mm(3) window range. Validation results showed that the proposed method was able to register CT lung datasets with TRE and ICE values <3 mm. In addition, the average number of voxel that failed the Gamma criteria was <3 %, which supports the clinical applicability of the propose registration mechanism. The proposed 4D-CT registration computes the volumetric lung deformations within clinically viable accuracy.

  18. Diagnostic performance of contrast enhanced CT and 18F-FDG PET/CT in suspicious recurrence of biliary tract cancer after curative resection

    International Nuclear Information System (INIS)

    Lee, Yun-Gyoo; Bang, Yung-Jue; Han, Sae-Won; Oh, Do-Youn; Chie, Eui Kyu; Jang, Jin-Young; Im, Seock-Ah; Kim, Tae-You; Kim, Sun-Whe; Ha, Sung Whan

    2011-01-01

    Because of the late clinical presentation of biliary tract cancer (BTC), only 10% of patients are eligible for curative surgery. Even among those patients who have undergone curative surgery, most patients develop recurrent cancer. This study is to determine the clinical role of 18 F-FDG PET/CT during post-operative surveillance of suspected recurrent BTC based on symptoms, laboratory findings and contrast-enhanced CT (ceCT) findings. We consecutively enrolled 50 patients with BTC who underwent curative surgery. An 18 F-FDG PET/CT was obtained for assessment of recurrence based on clinical suspicion during post-operative surveillance. The final confirmation of recurrence was determined pathologically or clinically. When a pathologic confirmation was impossible or inconclusive, a clinical confirmation was used by radiologic correlation with subsequent follow-up ceCT at a minimum of 3-month intervals. Diagnostic efficacy was evaluated by comparing the results of ceCT and 18 F-FDG PET/CT with the final diagnosis. Among the 50 patients, 34(68%) were confirmed to have a recurrence. PET/CT showed higher sensitivity (88% vs. 76%, p = 0.16) and accuracy (82% vs. 66%, p = 0.11) for recurrence compared to ceCT, even though the difference was not significant. The positive (86% vs. 74%, p = 0.72) and negative predictive values for recurrence (73% vs. 47%, p = 0.55) were not significantly different between PET/CT and ceCT. However, an additional PET/CT on ceCT significantly improved the sensitivity than did a ceCT alone (94% [32/34] for PET/CT on ceCT vs. 76% [26/34] for ceCT alone, p = 0.03) without increasing the specificity, positive predictive value, and negative predictive value. 18 F-FDG PET/CT alone is not more sensitive or specific than ceCT in the detection of recurrent BTC after curative surgery. These results do not reach statistical significance, probably due to the low number of patients. However, an additional 18 F-FDG PET/CT on ceCT significantly improves the

  19. MRI-based treatment planning with pseudo CT generated through atlas registration.

    Science.gov (United States)

    Uh, Jinsoo; Merchant, Thomas E; Li, Yimei; Li, Xingyu; Hua, Chiaho

    2014-05-01

    To evaluate the feasibility and accuracy of magnetic resonance imaging (MRI)-based treatment planning using pseudo CTs generated through atlas registration. A pseudo CT, providing electron density information for dose calculation, was generated by deforming atlas CT images previously acquired on other patients. The authors tested 4 schemes of synthesizing a pseudo CT from single or multiple deformed atlas images: use of a single arbitrarily selected atlas, arithmetic mean process using 6 atlases, and pattern recognition with Gaussian process (PRGP) using 6 or 12 atlases. The required deformation for atlas CT images was derived from a nonlinear registration of conjugated atlas MR images to that of the patient of interest. The contrasts of atlas MR images were adjusted by histogram matching to reduce the effect of different sets of acquisition parameters. For comparison, the authors also tested a simple scheme assigning the Hounsfield unit of water to the entire patient volume. All pseudo CT generating schemes were applied to 14 patients with common pediatric brain tumors. The image similarity of real patient-specific CT and pseudo CTs constructed by different schemes was compared. Differences in computation times were also calculated. The real CT in the treatment planning system was replaced with the pseudo CT, and the dose distribution was recalculated to determine the difference. The atlas approach generally performed better than assigning a bulk CT number to the entire patient volume. Comparing atlas-based schemes, those using multiple atlases outperformed the single atlas scheme. For multiple atlas schemes, the pseudo CTs were similar to the real CTs (correlation coefficient, 0.787-0.819). The calculated dose distribution was in close agreement with the original dose. Nearly the entire patient volume (98.3%-98.7%) satisfied the criteria of chi-evaluation (pediatric brain tumor patients. The doses calculated from pseudo CTs agreed well with those from real CTs

  20. Rectal cancer staging: Multidetector-row computed tomography diagnostic accuracy in assessment of mesorectal fascia invasion

    Science.gov (United States)

    Ippolito, Davide; Drago, Silvia Girolama; Franzesi, Cammillo Talei; Fior, Davide; Sironi, Sandro

    2016-01-01

    AIM: To assess the diagnostic accuracy of multidetector-row computed tomography (MDCT) as compared with conventional magnetic resonance imaging (MRI), in identifying mesorectal fascia (MRF) invasion in rectal cancer patients. METHODS: Ninety-one patients with biopsy proven rectal adenocarcinoma referred for thoracic and abdominal CT staging were enrolled in this study. The contrast-enhanced MDCT scans were performed on a 256 row scanner (ICT, Philips) with the following acquisition parameters: tube voltage 120 KV, tube current 150-300 mAs. Imaging data were reviewed as axial and as multiplanar reconstructions (MPRs) images along the rectal tumor axis. MRI study, performed on 1.5 T with dedicated phased array multicoil, included multiplanar T2 and axial T1 sequences and diffusion weighted images (DWI). Axial and MPR CT images independently were compared to MRI and MRF involvement was determined. Diagnostic accuracy of both modalities was compared and statistically analyzed. RESULTS: According to MRI, the MRF was involved in 51 patients and not involved in 40 patients. DWI allowed to recognize the tumor as a focal mass with high signal intensity on high b-value images, compared with the signal of the normal adjacent rectal wall or with the lower tissue signal intensity background. The number of patients correctly staged by the native axial CT images was 71 out of 91 (41 with involved MRF; 30 with not involved MRF), while by using the MPR 80 patients were correctly staged (45 with involved MRF; 35 with not involved MRF). Local tumor staging suggested by MDCT agreed with those of MRI, obtaining for CT axial images sensitivity and specificity of 80.4% and 75%, positive predictive value (PPV) 80.4%, negative predictive value (NPV) 75% and accuracy 78%; while performing MPR the sensitivity and specificity increased to 88% and 87.5%, PPV was 90%, NPV 85.36% and accuracy 88%. MPR images showed higher diagnostic accuracy, in terms of MRF involvement, than native axial images

  1. Enabling image fusion for a CT guided needle placement robot

    Science.gov (United States)

    Seifabadi, Reza; Xu, Sheng; Aalamifar, Fereshteh; Velusamy, Gnanasekar; Puhazhendi, Kaliyappan; Wood, Bradford J.

    2017-03-01

    Purpose: This study presents development and integration of hardware and software that enables ultrasound (US) and computer tomography (CT) fusion for a FDA-approved CT-guided needle placement robot. Having real-time US image registered to a priori-taken intraoperative CT image provides more anatomic information during needle insertion, in order to target hard-to-see lesions or avoid critical structures invisible to CT, track target motion, and to better monitor ablation treatment zone in relation to the tumor location. Method: A passive encoded mechanical arm is developed for the robot in order to hold and track an abdominal US transducer. This 4 degrees of freedom (DOF) arm is designed to attach to the robot end-effector. The arm is locked by default and is released by a press of button. The arm is designed such that the needle is always in plane with US image. The articulated arm is calibrated to improve its accuracy. Custom designed software (OncoNav, NIH) was developed to fuse real-time US image to a priori-taken CT. Results: The accuracy of the end effector before and after passive arm calibration was 7.07mm +/- 4.14mm and 1.74mm +/-1.60mm, respectively. The accuracy of the US image to the arm calibration was 5mm. The feasibility of US-CT fusion using the proposed hardware and software was demonstrated in an abdominal commercial phantom. Conclusions: Calibration significantly improved the accuracy of the arm in US image tracking. Fusion of US to CT using the proposed hardware and software was feasible.

  2. MRI-based treatment planning with pseudo CT generated through atlas registration

    International Nuclear Information System (INIS)

    Uh, Jinsoo; Merchant, Thomas E.; Hua, Chiaho; Li, Yimei; Li, Xingyu

    2014-01-01

    Purpose: To evaluate the feasibility and accuracy of magnetic resonance imaging (MRI)-based treatment planning using pseudo CTs generated through atlas registration. Methods: A pseudo CT, providing electron density information for dose calculation, was generated by deforming atlas CT images previously acquired on other patients. The authors tested 4 schemes of synthesizing a pseudo CT from single or multiple deformed atlas images: use of a single arbitrarily selected atlas, arithmetic mean process using 6 atlases, and pattern recognition with Gaussian process (PRGP) using 6 or 12 atlases. The required deformation for atlas CT images was derived from a nonlinear registration of conjugated atlas MR images to that of the patient of interest. The contrasts of atlas MR images were adjusted by histogram matching to reduce the effect of different sets of acquisition parameters. For comparison, the authors also tested a simple scheme assigning the Hounsfield unit of water to the entire patient volume. All pseudo CT generating schemes were applied to 14 patients with common pediatric brain tumors. The image similarity of real patient-specific CT and pseudo CTs constructed by different schemes was compared. Differences in computation times were also calculated. The real CT in the treatment planning system was replaced with the pseudo CT, and the dose distribution was recalculated to determine the difference. Results: The atlas approach generally performed better than assigning a bulk CT number to the entire patient volume. Comparing atlas-based schemes, those using multiple atlases outperformed the single atlas scheme. For multiple atlas schemes, the pseudo CTs were similar to the real CTs (correlation coefficient, 0.787–0.819). The calculated dose distribution was in close agreement with the original dose. Nearly the entire patient volume (98.3%–98.7%) satisfied the criteria of chi-evaluation (<2% maximum dose and 2 mm range). The dose to 95% of the volume and the

  3. MRI-based treatment planning with pseudo CT generated through atlas registration

    Energy Technology Data Exchange (ETDEWEB)

    Uh, Jinsoo, E-mail: jinsoo.uh@stjude.org; Merchant, Thomas E.; Hua, Chiaho [Department of Radiological Sciences, St. Jude Children' s Research Hospital, Memphis, Tennessee 38105 (United States); Li, Yimei; Li, Xingyu [Department of Biostatistics, St. Jude Children' s Research Hospital, Memphis, Tennessee 38105 (United States)

    2014-05-15

    Purpose: To evaluate the feasibility and accuracy of magnetic resonance imaging (MRI)-based treatment planning using pseudo CTs generated through atlas registration. Methods: A pseudo CT, providing electron density information for dose calculation, was generated by deforming atlas CT images previously acquired on other patients. The authors tested 4 schemes of synthesizing a pseudo CT from single or multiple deformed atlas images: use of a single arbitrarily selected atlas, arithmetic mean process using 6 atlases, and pattern recognition with Gaussian process (PRGP) using 6 or 12 atlases. The required deformation for atlas CT images was derived from a nonlinear registration of conjugated atlas MR images to that of the patient of interest. The contrasts of atlas MR images were adjusted by histogram matching to reduce the effect of different sets of acquisition parameters. For comparison, the authors also tested a simple scheme assigning the Hounsfield unit of water to the entire patient volume. All pseudo CT generating schemes were applied to 14 patients with common pediatric brain tumors. The image similarity of real patient-specific CT and pseudo CTs constructed by different schemes was compared. Differences in computation times were also calculated. The real CT in the treatment planning system was replaced with the pseudo CT, and the dose distribution was recalculated to determine the difference. Results: The atlas approach generally performed better than assigning a bulk CT number to the entire patient volume. Comparing atlas-based schemes, those using multiple atlases outperformed the single atlas scheme. For multiple atlas schemes, the pseudo CTs were similar to the real CTs (correlation coefficient, 0.787–0.819). The calculated dose distribution was in close agreement with the original dose. Nearly the entire patient volume (98.3%–98.7%) satisfied the criteria of chi-evaluation (<2% maximum dose and 2 mm range). The dose to 95% of the volume and the

  4. Intraoperative CT with integrated navigation system in spinal neurosurgery

    International Nuclear Information System (INIS)

    Zausinger, S.; Heigl, T.; Scheder, B.; Schnell, O.; Tonn, J.C.; Uhl, E.; Morhard, D.

    2007-01-01

    For spinal surgery navigational system images are usually acquired before surgery with patients positioned supine. The aim of this study was to evaluate prospectively navigated procedures in spinal surgery with data acquisition by intraoperative computed tomography (iCT). CT data of 38 patients [thoracolumbar instability (n = 24), C1/2 instability (n = 6), cervicothoracic stabilization (n = 7), disk herniation (n = 1)] were acquired after positioning the patient in prone position. A sliding gantry 24 detector row CT was used for image acquisition. Data were imported to the frameless infrared-based neuronavigation station. A postprocedural CT was obtained to assess the extent of decompression and the accuracy of instrumentation. Intraoperative registration revealed computed accuracy 2 mm in 9/158 screws (5.6%), allowing immediate correction in five screws without any damage to vessels or nerves. There were three transient complications with clinical improvement in all patients. Intraoperative CT in combination with neuronavigation provides high accuracy of screw placement and thus safety for patients undergoing spinal stabilization. The procedure is rapid and easy to perform and - by replacing pre- and postoperative imaging-is not associated with additional exposure to radiation. (orig.)

  5. The Added Diagnostic Value of 18F-Fluorodihydroxyphenylalanine PET/CT in the Preoperative Work-Up of Small Bowel Neuroendocrine Tumors.

    Science.gov (United States)

    Addeo, Pietro; Poncet, Gilles; Goichot, Bernard; Leclerc, Loic; Brigand, Cécile; Mutter, Didier; Romain, Benoit; Namer, Izzie-Jacques; Bachellier, Philippe; Imperiale, Alessio

    2018-04-01

    The precise localization of the primary tumor and/or the identification of multiple primary tumors improves the preoperative work-up in patients with small bowel (SB) neuroendocrine tumor (NET). The present study assesses the diagnostic value of 18 F-fluorodihydroxyphenylalanine ( 18 F-FDOPA) positron emission tomography/computed tomography (PET/CT) during the preoperative wok-up of SB NETs. Between January 2010 and June 2017, all consecutive patients with SB NETs undergoing preoperative 18 F-FDOPA PET/CT and successive resection were analyzed. Preoperative work-up included computed tomography (CT), somatostatin receptor scintigraphy (SRS), and 18 F-FDOPA PET/CT. Sensitivity and accuracy ratio for primary and multiple tumor detection were compared with data from surgery and pathology. There were 17 consecutive patients with SB NETs undergoing surgery. Nine patients (53%) had multiple tumors, 15 (88%) metastatic lymph nodes, 3 (18%) peritoneal carcinomatosis, and 9 patients (53%) liver metastases. A total of 70 SB NETs were found by pathology. Surgery identified the primary in 17/17 (100%) patients and recognized seven of 9 patients (78%) with multiple synchronous SB. Preoperatively, 18 F-FDOPA PET/CT displayed a statistically significant higher sensitivity for primary tumor localization (100 vs. 23.5 vs. 29.5%) and multiple tumor detection (78 vs. 22 vs. 11%) over SRS and CT. Compared with pathology, 18 F-FDOPA PET/CT displayed the highest accuracy ratio for number of tumor detected over CT and SRS (2.0 ± 2.2 vs. 0.4 ± 0.7 vs. 0.6 ± 1.5, p = 0.0003). 18 F-FDOPA PET/CT significantly increased the sensitivity and accuracy for primary and multiple SB NET identification. 18 F-FDOPA PET/CT should be included systematically in the preoperative work-up of SB NET.

  6. Influence of heart rate on diagnostic accuracy and image quality of 16-slice CT coronary angiography: comparison of multisegment and halfscan reconstruction approaches

    International Nuclear Information System (INIS)

    Dewey, Marc; Teige, Florian; Hamm, Bernd; Laule, Michael

    2007-01-01

    The lower the heart rate the better image quality in multislice computed tomography (MSCT) coronary angiography. We prospectively assessed the influence of heart rate on per-patient diagnostic accuracy and image quality of MSCT coronary angiography and compared adaptive multisegment and standard halfscan reconstruction. A consecutive cohort of 126 patients scheduled to undergo conventional coronary angiography was examined with 16-slice CT. For all heart rate groups, per-patient diagnostic accuracy was significantly higher for multisegment than halfscan reconstruction with values of 95 vs. 79% (p 74 bpm, 41 patients). Differences in diagnostic accuracy between adjacent heart rate groups were only significant for halfscan reconstruction for the comparison between the 65-74 and >74 bpm group (p < 0.05). The vessel lengths free of motion artifacts were significantly longer with multisegment reconstruction in all heart rate groups and for all coronary arteries (p < 0.005). For noninvasive MSCT coronary angiography, both per-patient diagnostic accuracy and image quality decline with increasing heart rate, and multisegment reconstruction at high heart rates yields similar results as standard halfscan reconstruction at low heart rates. (orig.)

  7. SU-F-P-32: A Phantom Study of Accuracy of Four-Dimensional Cone-Beam CT (4D-CBCT) Vs. Three-Dimensional Cone Beam CT (3D-CBCT) in Image Guided Radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    He, R; Morris, B; Duggar, N; Markovich, A; Standford, J; Lu, J; Yang, C [University of Mississippi Med. Center, Jackson, MS (United States)

    2016-06-15

    Purpose: SymmetryTM 4D IGRT system of Elekta has been installed at our institution, which offers the 4D CBCT registration option. This study is to evaluate the accuracy of 4D CBCT system by using the CIRS 4D motion phantom and to perform a feasibility study on the implementation of 4D-CBCT as image guidance for SBRT treatment. Methods: The 3D and 4D CT image data sets are acquired using the CIRS motion phantom on a Philips large bore CT simulator. The motion was set as 0.5 cm superior and inferior directions with 6 seconds recycle time. The 4D CT data were sorted as 10 phases. One identifiable part of the 4D CT QA insert from CIRS phantom was used as the target. The ITV MIP was drawn based on maximum intensity projection (MIP) and transferred as a planning structure into 4D CBCT system. Then the 3D CBCT and 4D CBCT images were taken and registered with the free breath (3D), MIP (4D) and average intensity projection (AIP)(4D) reference data sets. The couch shifts (X, Y, Z) are recorded and compared. Results: Table 1 listed the twelve couch shifts based on the registration of MIP, AIP and free breath CT data sets with 3D CBCT and 4D CBCT for both whole body and local registration. X, Y and Z represent couch shifts in the direction of the right-left, superior-inferior and anterior-posterior. The biggest differences of 0.73 cm and 0.57 cm are noted in the free breath CT data with 4D CBCT and 3D CBCT data registration. Fig. 1 and Fig. 2 are the shift analysis in diagram. Fig. 3 shows the registration. Conclusion: Significant differences exist in the shifts corresponding with the direction of target motion. Further investigations are ongoing.

  8. Interleaving cerebral CT perfusion with neck CT angiography. Pt. I. Proof of concept and accuracy of cerebral perfusion values

    Energy Technology Data Exchange (ETDEWEB)

    Oei, Marcel T.H.; Meijer, Frederick J.A.; Woude, Willem-Jan van der; Smit, Ewoud J.; Ginneken, Bram van; Prokop, Mathias; Manniesing, Rashindra [Radboud University Medical Center, Department of Radiology and Nuclear Medicine, P.O. Box 9101, Nijmegen (Netherlands)

    2017-06-15

    We present a novel One-Step-Stroke protocol for wide-detector CT scanners that interleaves cerebral CTP with volumetric neck CTA (vCTA). We evaluate whether the resulting time gap in CTP affects the accuracy of CTP values. Cerebral CTP maps were retrospectively obtained from 20 patients with suspicion of acute ischemic stroke and served as the reference standard. To simulate a 4 s gap for interleaving CTP with vCTA, we eliminated one acquisition at various time points of CTP starting from the bolus-arrival-time(BAT). Optimal timing of the vCTA was evaluated. At the time point with least errors, we evaluated elimination of a second time point (6 s gap). Mean absolute percentage errors of all perfusion values remained below 10 % in all patients when eliminating any one time point in the CTP sequence starting from the BAT. Acquiring the vCTA 2 s after reaching a threshold of 70HU resulted in the lowest errors (mean <3.0 %). Eliminating a second time point still resulted in mean errors <3.5 %. CBF/CBV showed no significant differences in perfusion values except MTT. However, the percentage errors were always below 10 % compared to the original protocol. Interleaving cerebral CTP with neck CTA is feasible with minor effects on the perfusion values. (orig.)

  9. CT-Urography

    International Nuclear Information System (INIS)

    Dalla Palma, Ludovico; Grotto, Maurizio; Morra, Aldo

    2005-01-01

    In this paper we present an overview of CT-Urography. With the advent of multislice CT scanners and the evolution of image processing methods this technique now affords optimal urographic images comparable to those obtained with conventional techniques. We describe the acquisition techniques and protocols used by the various authors. Effective radiation dose has conditioned the use of CT-Urography so that the tendency today is to reduce the number of scans by performing, after the non enhanced scan, a single contrast-enhanced scan comprising both the nephrographic and urographic phase. With the use of multislice CT the quality of the urogram improves with the number of slices. We illustrate a variety of processing techniques, multiplanar reconstruction (MPR), maximum (MIP) and average intensity projection (AIP) and volume rendering (VR) and present a series of upper urinary tract tumours testifying to the superiority of the AIP technique over MIP. We then review the results of comparative studies of CT-Urography with conventional urography in upper urinary tract diagnostics. Finally, we describe the advantages and limitations of CT-Urography [it

  10. EEG channels reduction using PCA to increase XGBoost's accuracy for stroke detection

    Science.gov (United States)

    Fitriah, N.; Wijaya, S. K.; Fanany, M. I.; Badri, C.; Rezal, M.

    2017-07-01

    In Indonesia, based on the result of Basic Health Research 2013, the number of stroke patients had increased from 8.3 ‰ (2007) to 12.1 ‰ (2013). These days, some researchers are using electroencephalography (EEG) result as another option to detect the stroke disease besides CT Scan image as the gold standard. A previous study on the data of stroke and healthy patients in National Brain Center Hospital (RS PON) used Brain Symmetry Index (BSI), Delta-Alpha Ratio (DAR), and Delta-Theta-Alpha-Beta Ratio (DTABR) as the features for classification by an Extreme Learning Machine (ELM). The study got 85% accuracy with sensitivity above 86 % for acute ischemic stroke detection. Using EEG data means dealing with many data dimensions, and it can reduce the accuracy of classifier (the curse of dimensionality). Principal Component Analysis (PCA) could reduce dimensionality and computation cost without decreasing classification accuracy. XGBoost, as the scalable tree boosting classifier, can solve real-world scale problems (Higgs Boson and Allstate dataset) with using a minimal amount of resources. This paper reuses the same data from RS PON and features from previous research, preprocessed with PCA and classified with XGBoost, to increase the accuracy with fewer electrodes. The specific fewer electrodes improved the accuracy of stroke detection. Our future work will examine the other algorithm besides PCA to get higher accuracy with less number of channels.

  11. CT-guided biopsy of pulmonary lesions : a comparison of diagnostic accuracy and complication rate between automated gun biopsy and fine needle aspiration biopsy

    International Nuclear Information System (INIS)

    Kim, Keon Woo; Kang, Duk Sik

    1998-01-01

    The purpose of this study is to compare the efficacy and safety of CT-guided automated gun biopsy with those of fine needle aspiration biopsy of pulmonary lesions. Under CT guidance, we performed automated gu biopsies in 115 cases of 109 patients and fine needle aspiration biopsies in 119 cases of 108 patients with pulmonary lesions. Between the two methods, we compared the diagnostic rate, diagnostic accuracy and frequency of compilations according to the depth and diameter of pulmonary lesions. The overall diagnostic rates of automated gun biopsy and fine needle aspiration biopsy were 76.5 % (88/115) and 64.7 % (77/119) respectively. There was a significant statistical difference (p=0.048), especially in the case of malignant lesions less than 3 cm (p=0.027) and more than 6 cm in maximal diameter (p=0.008), and in the lesions located from 1 cm to less than 3 cm from the pleural (p=0.030), as seen on CT. There was no significant statistical difference in the frequency of complications. Automated gun biopsy of pulmonary lesions under CT guidance is safe, with complications rate comparable to those of fine needle aspiration biopsy. A higher overall diagnostic rate can be achieved by automated gun biopsy than by fine needle aspiration biopsy. For the diagnosis of pulmonary lesions under CT guidance, automated gun biopsy is therefore a more useful procedure than fine needle aspiration biopsy. (author). 24 refs., 3 tabs

  12. Diagnosis of preoperative staging of bladder cancer by CT and angiography

    International Nuclear Information System (INIS)

    Munechika, Hirotsugu; Saito, Kazuhiko; Tanaka, Osamu; Tomiie, Fumitaka; Aihara, Toshinori

    1982-01-01

    The accuracy of computed tomography (CT) and angiography in defining the extent of local invasion was studied retrospectively in 19 patients with surgically proven carcinoma of the bladder. The overall accuracy of CT and angiographic staging in these cases was 54 percent and 47 percent respectively. The diagnosis of carcinoma of the bladder by CT tended to do overstaging rather than understaging. A tangential view of the tumor-bearing area by the selective internal iliac arteriography allows an accurate estimation of the perivesical tumor extension. The topographic diagnosis should be made by CT prior to the angiographic study. (author)

  13. Usefulness and limit of CT diagnosis on appendicitis

    International Nuclear Information System (INIS)

    Kuchiki, Megumi; Takanashi, Toshiyasu; Yamaguchi, Koichi.

    1997-01-01

    CT was performed in 104 patients with abdominal pain suspected appendicitis. CT showed positive finding (abnormal appendix, appendicolith, pericecal inflammatory change, fluid collection, LN swelling, abscess) and complication of appendicitis clearly. CT diagnosis showed high accuracy than clinical diagnosis. CT proved its usefulness especially only CT imaging showed the correct diagnosis. On the other hand, diverticulitis and terminal ileitis common diagnostic disease of appendicitis showed similar clinical appearance and CT image, caused to be difficult to diagnose correctly. In the cases showing similar image to appendicitis or atypical image, CT also proved its limit of the diagnosis on appendicitis. (author)

  14. Multi-institutional MicroCT image comparison of image-guided small animal irradiators

    Science.gov (United States)

    Johnstone, Chris D.; Lindsay, Patricia; E Graves, Edward; Wong, Eugene; Perez, Jessica R.; Poirier, Yannick; Ben-Bouchta, Youssef; Kanesalingam, Thilakshan; Chen, Haijian; E Rubinstein, Ashley; Sheng, Ke; Bazalova-Carter, Magdalena

    2017-07-01

    To recommend imaging protocols and establish tolerance levels for microCT image quality assurance (QA) performed on conformal image-guided small animal irradiators. A fully automated QA software SAPA (small animal phantom analyzer) for image analysis of the commercial Shelley micro-CT MCTP 610 phantom was developed, in which quantitative analyses of CT number linearity, signal-to-noise ratio (SNR), uniformity and noise, geometric accuracy, spatial resolution by means of modulation transfer function (MTF), and CT contrast were performed. Phantom microCT scans from eleven institutions acquired with four image-guided small animal irradiator units (including the commercial PXi X-RAD SmART and Xstrahl SARRP systems) with varying parameters used for routine small animal imaging were analyzed. Multi-institutional data sets were compared using SAPA, based on which tolerance levels for each QA test were established and imaging protocols for QA were recommended. By analyzing microCT data from 11 institutions, we established image QA tolerance levels for all image quality tests. CT number linearity set to R 2  >  0.990 was acceptable in microCT data acquired at all but three institutions. Acceptable SNR  >  36 and noise levels  1.5 lp mm-1 for MTF  =  0.2) was obtained at all but four institutions due to their large image voxel size used (>0.275 mm). Ten of the eleven institutions passed the set QA tolerance for geometric accuracy (2000 HU for 30 mgI ml-1). We recommend performing imaging QA with 70 kVp, 1.5 mA, 120 s imaging time, 0.20 mm voxel size, and a frame rate of 5 fps for the PXi X-RAD SmART. For the Xstrahl SARRP, we recommend using 60 kVp, 1.0 mA, 240 s imaging time, 0.20 mm voxel size, and 6 fps. These imaging protocols should result in high quality images that pass the set tolerance levels on all systems. Average SAPA computation time for complete QA analysis for a 0.20 mm voxel, 400 slice Shelley phantom microCT data set

  15. SPECT/CT and pulmonary embolism

    DEFF Research Database (Denmark)

    Mortensen, Jann; Borgwardt, Henrik Gutte

    2014-01-01

    the best diagnostic accuracy for PE. In addition, recent developments in the CTPA technique have made it possible to image the pulmonary arteries of the lungs in one breath-hold. This development is based on the change from a single-detector to multidetector CT technology with an increase in volume......Acute pulmonary embolism (PE) is diagnosed either by ventilation/perfusion (V/P) scintigraphy or pulmonary CT angiography (CTPA). In recent years both techniques have improved. Many nuclear medicine centres have adopted the single photon emission CT (SPECT) technique as opposed to the planar...

  16. Accuracy evaluation of fusion of CT, MR, and SPECT images using commercially available software packages (SRS PLATO and IFS)

    International Nuclear Information System (INIS)

    Mongioj, Valeria; Brusa, Anna; Loi, Gianfranco; Pignoli, Emanuele; Gramaglia, Alberto; Scorsetti, Marta; Bombardieri, Emilio; Marchesini, Renato

    1999-01-01

    Purpose: A problem for clinicians is to mentally integrate information from multiple diagnostic sources, such as computed tomography (CT), magnetic resonance (MR), and single photon emission computed tomography (SPECT), whose images give anatomic and metabolic information. Methods and Materials: To combine this different imaging procedure information, and to overlay correspondent slices, we used commercially available software packages (SRS PLATO and IFS). The algorithms utilize a fiducial-based coordinate system (or frame) with 3 N-shaped markers, which allows coordinate transformation of a clinical examination data set (9 spots for each transaxial section) to a stereotactic coordinate system. The N-shaped markers were filled with fluids visible in each modality (gadolinium for MR, calcium chloride for CT, and 99m Tc for SPECT). The frame is relocatable, in the different acquisition modalities, by means of a head holder to which a face mask is fixed so as to immobilize the patient. Position errors due to the algorithms were obtained by evaluating the stereotactic coordinates of five sources detectable in each modality. Results: SPECT and MR position errors due to the algorithms were evaluated with respect to CT: Δx was ≤ 0.9 mm for MR and ≤ 1.4 mm for SPECT, Δy was ≤ 1 mm and ≤ 3 mm for MR and SPECT, respectively. Maximal differences in distance between estimated and actual fiducial centers (geometric mismatch) were in the order of the pixel size (0.8 mm for CT, 1.4 mm for MR, and 1.8 mm for SPECT). In an attempt to distinguish necrosis from residual disease, the image fusion protocol was studied in 35 primary or metastatic brain tumor patients. Conclusions: The image fusion technique has a good degree of accuracy as well as the potential to improve the specificity of tissue identification and the precision of the subsequent treatment planning

  17. CT Angiography of Peripheral Arterial Disease by 256-Slice Scanner: Accuracy, Advantages and Disadvantages Compared to Digital Subtraction Angiography.

    Science.gov (United States)

    Mishra, Atul; Jain, Narendra; Bhagwat, Anand

    2017-07-01

    Peripheral arterial occlusive disease (PAOD) may cause disabling claudication or critical limb ischemia. Multidetector computed tomography (CT) technology has evolved to the level of 256-slice CT scanners which has significantly improved the spatial and temporal resolution of the images. This has provided the capability of chasing the contrast bolus at a fast speed enabling angiographic imaging of long segments of the body. These images can be reconstructed in various planes and various modes for detailed analysis of the peripheral vascular diseases which helps in making treatment decision. The aim of this retrospective study was to compare the CT angiograms (CTAs) of all cases of PAOD done by 256-slice CT scanner at a tertiary care vascular center and comparing these images with the digital subtraction angiograms (DSAs) of these patients. The retrospective study included 53 patients who underwent both CTA and DSA at our center over a period of 3 years from March 2013 to March 2016. The CTA showed high sensitivity (93%) and specificity (92.7%) for overall assessment of degree of stenosis in a vascular segment in cases of aortic and lower limb occlusive disease. The assessment of lesions of infrapopliteal segment was comparatively inferior (sensitivity 91.6%, accuracy 73.3%, and positive predictive value 78.5%), more so in the presence of significant calcification. The advantages of CTA were its noninvasive nature, ability to image large area of body, almost no adverse effects to the patients, and better assessment of vessel wall disease. However, the CTA assessment of collaterals was inferior with a sensitivity of only 62.7% as compared to DSA. Overall, 256-slice CTA provides fast and accurate imaging of vascular tree which can restrict DSA only in few selected cases as a problem-solving tool where clinico-radiological mismatch is present.

  18. Brain tumor and CT, 1

    International Nuclear Information System (INIS)

    Suzuki, Nobuyuki; Katada, Kazuhiro; Shinomiya, Youichi; Sano, Hirotoshi; Kanno, Tetsuo

    1981-01-01

    It is very important for a neurosurgeon to know the consistency of a brain tumor preoperatively, since the information which is of much use in indicating the likely difficulty of the operation, which operative tools should be selected, the amount of bleeding to be expected from the tumor, and so on. The authors, therefore, tried to evaluate the consistency of brain tumors preoperatively 27 cases in which the margin of the tumor was made clear with a homogeneous stain were studied concerning the relationship between the tumor consistency and the CT findings. The results are as follows: 1) A higher CT number on a plain CT indicated a harder consistency of the tumor. 2) A lesser contrast index (CT number on enhancement CT/CT number on plain CT) showed a harder consistency of the tumor. (author)

  19. Value of surveillance {sup 18}F FDG PET/CT in colorectal cancer:comparison with conventional imaging studies

    Energy Technology Data Exchange (ETDEWEB)

    Choi, Eun Kyoung; Yoo, Ie Ryung; Park, Hye Lim; Choi, Hyun Su; Han, Eun Ji; Kim, Sung Hoon; Chung, Soo Kyo; O, Joo Hyun [The Catholic Univ. of Korea, Seoul (Korea, Republic of)

    2012-09-15

    To assess the value of PET/CT for detecting local or distant recurrence in patients who undergo surgery for colorectal cancer (CRC)and to compare the accuracy of PET/CT to that of conventional imaging studies (CIS). Tumor surveillance PET/CT scans done between March 2005 and December 2009 of disease free patients after surgery with or without adjuvant chemotherapy for CRC were retrospectively studied. CIS (serial enhanced CT from lung base to pelvis and plain chest radiograph)were performed within 1 month of PET/CT. We excluded patients with distant metastasis on initial staging, a known recurrent tumor, and a lack of follow up imaging. The final diagnosis was based on at least 6 months of follow up with colonoscopy, biopsy, and serial imaging studies in combination with carcinoembryonic antigen levels. A total of 262 PET/CT scans of 245 patients were included. Local and distant recurrences were detected in 27 cases (10.3%). On case based analysis, the overall sensitivity, specificity, and accuracy were 100, 97.0, and 97.3% for PET/CT and 85.1, 97.0, and 95.8% for CIS, respectively. On lesion based analysis, PET/CT detected more lesions compared to CIS in local recurrence and lung metastasis. PET/CT and CIS detected the same number of lesions in abdominal lymph nodes, hepatic metastasis, and peritoneal carcinomatosis. PET/CT detected two more metachronous tumors than did CIS in the lung and thyroid gland. PET/CT detected more recurrences in patients who underwent surgery for CRC than did CIS and had the additional advantage of evaluating the entire body during a single scan.

  20. Generating Evidence for Clinical Benefit of PET/CT in Diagnosing Cancer Patients

    DEFF Research Database (Denmark)

    Vach, Werner; Høilund-Carlsen, Poul Flemming; Gerke, Oke

    2011-01-01

    For diagnostic methods such as PET/CT, not only diagnostic accuracy but also clinical benefit must be demonstrated. However, there is a lack of consensus about how to approach this task. Here we consider 6 clinical scenarios to review some basic approaches to demonstrating the clinical benefit...... of PET/CT in cancer patients: replacement of an invasive procedure, improved accuracy of initial diagnosis, improved accuracy of staging for curative versus palliative treatment, improved accuracy of staging for radiation versus chemotherapy, response evaluation, and acceleration of clinical decisions....... We also develop some guidelines for the evaluation of clinical benefit. First, it should be clarified whether there is a direct benefit of the use of PET/CT or an indirect benefit because of improved diagnostic accuracy. If there is an indirect benefit, then decision modeling should be used initially...

  1. Recurrent bladder carcinoma: clinical and prognostic role of 18 F-FDG PET/CT

    Energy Technology Data Exchange (ETDEWEB)

    Alongi, Pierpaolo [San Raffaele G. Giglio Institute, Department of Radiological Sciences, Nuclear Medicine Unit, Cefalu (Italy); Caobelli, Federico [Basel University Hospital, Department of Nuclear Medicine, Basel (Switzerland); Gentile, Roberta; Baldari, Sergio [University of Messina, Nuclear Medicine Unit, Department of Biomedical Sciences and Morphological and Functional Images, Messina (Italy); Stefano, Alessandro; Russo, Giorgio; Gilardi, Maria Carla [IBFM-CNR, Cefalu (Italy); Albano, Domenico [Universita degli Studi di Palermo, DIBIMEF - Sezione di Scienze Radiologiche, Palermo (Italy); Midiri, Massimo [San Raffaele G. Giglio Institute, Department of Radiological Sciences, Nuclear Medicine Unit, Cefalu (Italy); Universita degli Studi di Palermo, DIBIMEF - Sezione di Scienze Radiologiche, Palermo (Italy)

    2017-02-15

    A small number of studies evaluated the detection rate of lesions from bladder carcinoma (BC) of 18 F-FDG PET/CT in the restaging process. However, the prognostic role of FDG PET/CT still remains unclear. The aim of the present study was to evaluate the accuracy, the effect upon treatment decision, and the prognostic value of FDG PET/CT in patients with suspected recurrent BC. Forty-one patients affected by BC underwent FDG PET/CT for restaging purpose. The diagnostic accuracy of visually interpreted FDG PET/CT was assessed compared to histology (n = 8), other diagnostic imaging modalities (contrast-enhanced CT in 38/41 patients and MRI in 15/41) and clinical follow-up (n = 41). Semiquantitative PET values (SUVmax, SUVmean, SUL, MTV, TLG) were calculated using a graph-based method. Progression-free survival (PFS) and overall survival (OS) were assessed by using Kaplan-Meier curves. The risk of progression (hazard ratio, HR) was computed by Cox regression analysis by considering all the available variables. PET was considered positive in 21 of 41 patients. Of these, recurrent BC was confirmed in 20 (95 %). Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy of FDG PET/CT were 87 %, 94 %, 95 %, 85 %, 90 %. AUC was 0.9 (95 %IC 0.8-1). Bayesian positive and negative likelihood ratios were 14.5 and 0.13, respectively. FDG PET/CT findings modified the therapeutic approach in 16 patients (modified therapy in 10 PET-positive patients, watch-and-wait in six PET-negative patients). PFS was significantly longer in patients with negative scan vs. those with pathological findings (85 % vs. 24 %, p < 0.05; HR = 12.4; p = 0.001). Moreover, an unremarkable study was associated with a longer OS (88 % vs. 47 % after 2 years and 87 % vs. 25 % after 3 years, respectively, p < 0.05). Standardized uptake value (SUV)max > 6 and total lesion glycolysis (TLG) > 8.5 were recognized as the most accurate thresholds to predict PFS (2-year PFS 62 % for

  2. Respiratory management of CT-transmission for accuracy fusion in PET/CT. A comparison between normal expiration and free breathing in 600 experiences

    International Nuclear Information System (INIS)

    Osawa, Atsushi; Takiguchi, Tomohiro; Tamura, Shintaro; Ohashi, Takashi; Miwa, Kenta; Akimoto, Kenta; Wagatsuma, Kei

    2010-01-01

    Image misregistration can occur in fusion positron emission tomography (PET)/CT, because of motion artifacts caused by the management of respiration. The standard imaging protocol of the CT component of PET/CT is normal expiration (NormExp) or free breathing (FB). The objective of this study was to compare NormExp and FB for the optimal breathing protocol for PET/CT scans. A total of 600 consecutive patients were examined using lutetium oxyorthosilicate (LSO)-based PET/CT. CT was acquired during NormExp (id est (i.e.), the level reached when the patient exhaled without forcing expiration and then held the breath) in 300 patients and during FB in 300 patients. The profile of liver measured along body axis was assessed. The distance of profile centers between the PET image and the CT image was measured. The misalignment between profile centers (PET) and profile centers (CT) was compared between NormExp and FB using the histogram of patients. An F test was used to test if the variances of two misalignments are equal. Next, the relationship between misalignment and age was evaluated in two managements of respiration. There was no significant difference between NormExp and FB in the histogram. However, significant misalignments (>10 cm) were found with NormExp. Patient age may have influenced the mismatch. FB is recommended for geriatric patients during acquisition of attenuation correction CT data sets. (author)

  3. PET/CT. Dose-escalated image fusion?

    International Nuclear Information System (INIS)

    Brix, G.; Beyer, T.

    2005-01-01

    Clinical studies demonstrate a gain in diagnostic accuracy by employing combined PET/CT instead of separate CT and PET imaging. However, whole-body PET/CT examinations result in a comparatively high radiation burden to patients and thus require a proper justification and optimization to avoid repeated exposure or over-exposure of patients. This review article summarizes relevant data concerning radiation exposure of patients resulting from the different components of a combined PET/CT examination and presents different imaging strategies that can help to balance the diagnostic needs and the radiation protection requirements. In addition various dose reduction measures are discussed, some of which can be adopted from CT practice, while others mandate modifications to the existing hard- and software of PET/CT systems. (orig.)

  4. Philips Gemini TF64 PET/CT Acceptance Testing

    International Nuclear Information System (INIS)

    González Gonzalez, Joaquín J.; Calderón Marin, Carlos F.; Varela Corona, Consuelo; Machado Tejeda, Adalberto; González Correa, Héctor J.

    2016-01-01

    Na. Results. The manufacturer protocol for CT scanner passed all performed tests for head and body phantom such as image artifacts, mean CT number, CT uniformity, image noise, spatial resolution and slices thickness. The performed geometrical tests were within specifications. For PET scanner, the energy resolution and the system coincidence timing resolution was 12.03 % and 516.2 ps respectively. The measured values of transversal and axial resolutions near the center, absolute sensitivity at the center and at 10 cm off-center, scatter fraction, counting rate performance, accuracy for correction of count losses and random as well as image quality measurements of lesions contrast and background variability were within manufacturer specifications. A modest improvement of lesion detectability was observed when quality image phantom was reconstructed using TOF capability. The measured (X, Y, Z, Z rotation) offset values in mm of PET/CT images alignment with gantry in closed and open positions were (-3.22, 3.46, 1.61, -0.44) and (-1.2, 3.25, 3.28, -0.58) respectively. Conclusion. The measurement of performance characteristics of the Philips Gemini TF64 PET/CT scanner satisfies the manufacturer specification.

  5. Investigating CT to CBCT image registration for head and neck proton therapy as a tool for daily dose recalculation

    Energy Technology Data Exchange (ETDEWEB)

    Landry, Guillaume, E-mail: g.landry@lmu.de [Department of Medical Physics, Ludwig-Maximilians-University, Munich D85748, Germany and Department of Radiation Oncology, Ludwig-Maximilians-University, Munich D81377 (Germany); Nijhuis, Reinoud; Thieke, Christian; Reiner, Michael; Ganswindt, Ute; Belka, Claus [Department of Radiation Oncology, Ludwig-Maximilians-University, Munich D81377 (Germany); Dedes, George; Handrack, Josefine; Parodi, Katia [Department of Medical Physics, Ludwig-Maximilians-University, Munich D85748 (Germany); Janssens, Guillaume; Orban de Xivry, Jonathan [ICTEAM, Université Catholique de Louvain, Louvain-La-Neuve B1348 (Belgium); Kamp, Florian; Wilkens, Jan J. [Department of Radiation Oncology, Technische Universität München, Klinikum rechts der Isar, Munich D81675, Germany and Physik-Department, Technische Universität München, Garching D85748 (Germany); Paganelli, Chiara; Riboldi, Marco; Baroni, Guido [Dipartimento di Elettronica Informazione e Bioingegneria, Politecnico di Milano, Milan 20133 (Italy)

    2015-03-15

    Purpose: Intensity modulated proton therapy (IMPT) of head and neck (H and N) cancer patients may be improved by plan adaptation. The decision to adapt the treatment plan based on a dose recalculation on the current anatomy requires a diagnostic quality computed tomography (CT) scan of the patient. As gantry-mounted cone beam CT (CBCT) scanners are currently being offered by vendors, they may offer daily or weekly updates of patient anatomy. CBCT image quality may not be sufficient for accurate proton dose calculation and it is likely necessary to perform CBCT CT number correction. In this work, the authors investigated deformable image registration (DIR) of the planning CT (pCT) to the CBCT to generate a virtual CT (vCT) to be used for proton dose recalculation. Methods: Datasets of six H and N cancer patients undergoing photon intensity modulated radiation therapy were used in this study to validate the vCT approach. Each dataset contained a CBCT acquired within 3 days of a replanning CT (rpCT), in addition to a pCT. The pCT and rpCT were delineated by a physician. A Morphons algorithm was employed in this work to perform DIR of the pCT to CBCT following a rigid registration of the two images. The contours from the pCT were deformed using the vector field resulting from DIR to yield a contoured vCT. The DIR accuracy was evaluated with a scale invariant feature transform (SIFT) algorithm comparing automatically identified matching features between vCT and CBCT. The rpCT was used as reference for evaluation of the vCT. The vCT and rpCT CT numbers were converted to stopping power ratio and the water equivalent thickness (WET) was calculated. IMPT dose distributions from treatment plans optimized on the pCT were recalculated with a Monte Carlo algorithm on the rpCT and vCT for comparison in terms of gamma index, dose volume histogram (DVH) statistics as well as proton range. The DIR generated contours on the vCT were compared to physician-drawn contours on the rpCT

  6. Graded compression ultrasonography and computed tomography in acute colonic diverticulitis: Meta-analysis of test accuracy

    International Nuclear Information System (INIS)

    Lameris, Wytze; Randen, Adrienne van; Bipat, Shandra; Stoker, Jaap; Bossuyt, Patrick M.M.; Boermeester, Marja A.

    2008-01-01

    The purpose was to investigate the diagnostic accuracy of graded compression ultrasonography (US) and computed tomography (CT) in diagnosing acute colonic diverticulitis (ACD) in suspected patients. We performed a systematic review and meta-analysis of the accuracy of CT and US in diagnosing ACD. Study quality was assessed with the QUADAS tool. Summary estimates of sensitivity and specificity were calculated using a bivariate random effects model. Six US studies evaluated 630 patients, and eight CT studies evaluated 684 patients. Overall, their quality was moderate. We did not identify meaningful sources of heterogeneity in the study results. Summary sensitivity estimates were 92% (95% CI: 80%-97%) for US versus 94% (95%CI: 87%-97%) for CT (p = 0.65). Summary specificity estimates were 90% (95%CI: 82%-95%) for US versus 99% (95%CI: 90%-100%) for CT (p = 0.07). For the identification of alternative diseases sensitivity ranged between 33% and 78% for US and between 50% and 100% for CT. The currently best available evidence shows no statistically significant difference in accuracy of US and CT in diagnosing ACD. Therefore, both US and CT can be used as initial diagnostic tool until new evidence is brought forward. However, CT is more likely to identify alternative diseases. (orig.)

  7. Accuracy, calibration and clinical performance of the EuroSCORE: can we reduce the number of variables?

    Science.gov (United States)

    Ranucci, Marco; Castelvecchio, Serenella; Menicanti, Lorenzo; Frigiola, Alessandro; Pelissero, Gabriele

    2010-03-01

    The European system for cardiac operative risk evaluation (EuroSCORE) is currently used in many institutions and is considered a reference tool in many countries. We hypothesised that too many variables were included in the EuroSCORE using limited patient series. We tested different models using a limited number of variables. A total of 11150 adult patients undergoing cardiac operations at our institution (2001-2007) were retrospectively analysed. The 17 risk factors composing the EuroSCORE were separately analysed and ranked for accuracy of prediction of hospital mortality. Seventeen models were created by progressively including one factor at a time. The models were compared for accuracy with a receiver operating characteristics (ROC) analysis and area under the curve (AUC) evaluation. Calibration was tested with Hosmer-Lemeshow statistics. Clinical performance was assessed by comparing the predicted with the observed mortality rates. The best accuracy (AUC 0.76) was obtained using a model including only age, left ventricular ejection fraction, serum creatinine, emergency operation and non-isolated coronary operation. The EuroSCORE AUC (0.75) was not significantly different. Calibration and clinical performance were better in the five-factor model than in the EuroSCORE. Only in high-risk patients were 12 factors needed to achieve a good performance. Including many factors in multivariable logistic models increases the risk for overfitting, multicollinearity and human error. A five-factor model offers the same level of accuracy but demonstrated better calibration and clinical performance. Models with a limited number of factors may work better than complex models when applied to a limited number of patients. Copyright (c) 2009 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.

  8. Clinical PET/CT imaging. Promises and misconceptions

    International Nuclear Information System (INIS)

    Czernin, J.; Auerbach, M.A.

    2005-01-01

    PET/CT is now established as the most important imaging tool in oncology. PET/CT stages and restages cancer with a higher accuracy than PET or CT alone. The sometimes irrational approach to combine state of the art PET with the highest end CT devices should give way to a more reasonable equipment design tailored towards the specific clinical indications in well-defined patient populations. The continuing success of molecular PET/CT now depends more upon advances in molecular imaging with the introduction of targeted imaging probes for individualized therapy approaches in cancer patients and less upon technological advances of imaging equipment. (orig.)

  9. Dosimetric impact of a CT metal artefact suppression algorithm for proton, electron and photon therapies

    International Nuclear Information System (INIS)

    Wei Jikun; Sandison, George A; Hsi, W-C; Ringor, Michael; Lu Xiaoyi

    2006-01-01

    Accurate dose calculation is essential to precision radiation treatment planning and this accuracy depends upon anatomic and tissue electron density information. Modern treatment planning inhomogeneity corrections use x-ray CT images and calibrated scales of tissue CT number to electron density to provide this information. The presence of metal in the volume scanned by an x-ray CT scanner causes metal induced image artefacts that influence CT numbers and thereby introduce errors in the radiation dose distribution calculated. This paper investigates the dosimetric improvement achieved by a previously proposed x-ray CT metal artefact suppression technique when the suppressed images of a patient with bilateral hip prostheses are used in commercial treatment planning systems for proton, electron or photon therapies. For all these beam types, this clinical image and treatment planning study reveals that the target may be severely underdosed if a metal artefact-contaminated image is used for dose calculations instead of the artefact suppressed one. Of the three beam types studied, the metal artefact suppression is most important for proton therapy dose calculations, intermediate for electron therapy and least important for x-ray therapy but still significant. The study of a water phantom having a metal rod simulating a hip prosthesis indicates that CT numbers generated after image processing for metal artefact suppression are accurate and thus dose calculations based on the metal artefact suppressed images will be of high fidelity

  10. SU-F-J-214: Dose Reduction by Spatially Optimized Image Quality Via Fluence Modulated Proton CT (FMpCT)

    International Nuclear Information System (INIS)

    De Angelis, L; Landry, G; Dedes, G; Parodi, K; Hansen, D; Rit, S; Belka, C

    2016-01-01

    Purpose: Proton CT (pCT) is a promising imaging modality for reducing range uncertainty in image-guided proton therapy. Range uncertainties partially originate from X-ray CT number conversion to stopping power ratio (SPR) and are limiting the exploitation of the full potential of proton therapy. In this study we explore the concept of spatially dependent fluence modulated proton CT (FMpCT), for achieving optimal image quality in a clinical region of interest (ROI), while reducing significantly the imaging dose to the patient. Methods: The study was based on simulated ideal pCT using pencil beam (PB) scanning. A set of 250 MeV protons PBs was used to create 360 projections of a cylindrical water phantom and a head and neck cancer patient. The tomographic images were reconstructed using a filtered backprojection (FBP) as well as an iterative algorithm (ITR). Different fluence modulation levels were investigated and their impact on the image was quantified in terms of SPR accuracy as well as noise within and outside selected ROIs, as a function of imaging dose. The unmodulated image served as reference. Results: Both FBP reconstruction and ITR without total variation (TV) yielded image quality in the ROIs similar to the reference images, for modulation down to 0.1 of the full proton fluence. The average dose was reduced by 75% for the water phantom and by 40% for the patient. FMpCT does not improve the noise for ITR with TV and modulation 0.1. Conclusion: This is the first work proposing and investigating FMpCT for producing optimal image quality for treatment planning and image guidance, while simultaneously reducing imaging dose. Future work will address spatial resolution effects and the impact of FMpCT on the quality of proton treatment plans for a prototype pCT scanner capable of list mode data acquisition. Acknowledgement: DFG-MAP DFG - Munich-Centre for Advanced Photonics (MAP)

  11. SU-F-J-214: Dose Reduction by Spatially Optimized Image Quality Via Fluence Modulated Proton CT (FMpCT)

    Energy Technology Data Exchange (ETDEWEB)

    De Angelis, L; Landry, G; Dedes, G; Parodi, K [Ludwig-Maximilians-Universitaet Muenchen (LMU Munich), Garching b. Muenchen (Germany); Hansen, D [Aarhus University Hospital, Aarhus, Jutland (Denmark); Rit, S [University Lyon, Lyon, Auvergne-Rhone-Alpes (France); Belka, C [LMU Munich, Munich (Germany)

    2016-06-15

    Purpose: Proton CT (pCT) is a promising imaging modality for reducing range uncertainty in image-guided proton therapy. Range uncertainties partially originate from X-ray CT number conversion to stopping power ratio (SPR) and are limiting the exploitation of the full potential of proton therapy. In this study we explore the concept of spatially dependent fluence modulated proton CT (FMpCT), for achieving optimal image quality in a clinical region of interest (ROI), while reducing significantly the imaging dose to the patient. Methods: The study was based on simulated ideal pCT using pencil beam (PB) scanning. A set of 250 MeV protons PBs was used to create 360 projections of a cylindrical water phantom and a head and neck cancer patient. The tomographic images were reconstructed using a filtered backprojection (FBP) as well as an iterative algorithm (ITR). Different fluence modulation levels were investigated and their impact on the image was quantified in terms of SPR accuracy as well as noise within and outside selected ROIs, as a function of imaging dose. The unmodulated image served as reference. Results: Both FBP reconstruction and ITR without total variation (TV) yielded image quality in the ROIs similar to the reference images, for modulation down to 0.1 of the full proton fluence. The average dose was reduced by 75% for the water phantom and by 40% for the patient. FMpCT does not improve the noise for ITR with TV and modulation 0.1. Conclusion: This is the first work proposing and investigating FMpCT for producing optimal image quality for treatment planning and image guidance, while simultaneously reducing imaging dose. Future work will address spatial resolution effects and the impact of FMpCT on the quality of proton treatment plans for a prototype pCT scanner capable of list mode data acquisition. Acknowledgement: DFG-MAP DFG - Munich-Centre for Advanced Photonics (MAP)

  12. TU-H-207A-03: CT Hounsfield Unit Accuracy: Effect of Beam Hardening On Phantom and Clinical Whole-Body CT Images

    Energy Technology Data Exchange (ETDEWEB)

    Ai, H; Wendt, R [The University of Texas MD Anderson Cancer Center, Houston, TX (United States)

    2016-06-15

    Purpose: To assess the effect of beam hardening on measured CT HU values. Methods: An anthropomorphic knee phantom was scanned with the CT component of a GE Discovery 690 PET/CT scanner (120kVp, 300mAs, 40?0.625mm collimation, pitch=0.984, FOV=500mm, matrix=512?512) with four different scan setups, each of which induces different degrees of beam hardening by introducing additional attenuation media into the field of view. Homogeneous voxels representing “soft tissue” and “bone” were segmented by HU thresholding followed by a 3D morphological erosion operation which removes the non-homogenous voxels located on the interface of thresholded tissue mask. HU values of segmented “soft tissue” and “bone” were compared.Additionally, whole-body CT data with coverage from the skull apex to the end of toes were retrospectively retrieved from seven PET/CT exams to evaluate the effect of beam hardening in vivo. Homogeneous bone voxels were segmented with the same method previously described. Total In-Slice Attenuation (TISA) for each CT slice, defined as the summation of HU values over all voxels within a CT slice, was calculated for all slices of the seven whole-body CT datasets and evaluated against the mean HU values of homogeneous bone voxels within that slice. Results: HU values measured from the phantom showed that while “soft tissue” HU values were unaffected, added attenuation within the FOV caused noticeable decreases in the measured HU values of “bone” voxels. A linear relationship was observed between bone HU and TISA for slices of the torso and legs, but not of the skull. Conclusion: Beam hardening effect is not an issue of concern for voxels with HU in the soft tissue range, but should not be neglected for bone voxels. A linear relationship exists between bone HU and the associated TISA in non-skull CT slices, which can be exploited to develop a correction strategy.

  13. CT detection of thyroid pyramidal lobe in preoperative patients with thyroid tumors

    International Nuclear Information System (INIS)

    Shin, Gi Won; Kim, Dong Won

    2013-01-01

    Thyroid pyramidal lobe (TPL) is a normal variant of the thyroid gland, but few imaging studies of TPL have been published. The purpose of this study is to investigate the frequency, location, size (length, maximal AP diameter, maximal transverse diameter), and upper end level of TPL with its separation from the main thyroid gland on preoperative neck CT and to compare them with operative findings in order to assess the diagnostic accuracy of neck CT for detection TPL. 46 patients, who underwent preoperative neck CT before thyroidectomy, were included in the study. The frequency, location, size, and upper end level of TPL with its separation from the main thyroid gland on the neck CT was analyzed by a single radiologist. The sensitivity, specificity, positive predictive value, negative predictive value and accuracy of neck CT for detecting TPL was 77.8%, 89.5%, 91.3%, 73.9% and 82.6%. There was a significant difference in maximal AP diameter, location, upper end level, and its separation from main thyroid gland between CT and operative findings (p 0.05). Diagnostic accuracy of neck CT for detecting TPL was high, and the neck CT may be useful for evaluating TPL in the suprahyoid neck.

  14. Accuracy of estimation of graft size for living-related liver transplantation: first results of a semi-automated interactive software for CT-volumetry.

    Directory of Open Access Journals (Sweden)

    Theresa Mokry

    Full Text Available To evaluate accuracy of estimated graft size for living-related liver transplantation using a semi-automated interactive software for CT-volumetry.Sixteen donors for living-related liver transplantation (11 male; mean age: 38.2±9.6 years underwent contrast-enhanced CT prior to graft removal. CT-volumetry was performed using a semi-automated interactive software (P, and compared with a manual commercial software (TR. For P, liver volumes were provided either with or without vessels. For TR, liver volumes were provided always with vessels. Intraoperative weight served as reference standard. Major study goals included analyses of volumes using absolute numbers, linear regression analyses and inter-observer agreements. Minor study goals included the description of the software workflow: degree of manual correction, speed for completion, and overall intuitiveness using five-point Likert scales: 1--markedly lower/faster/higher for P compared with TR, 2--slightly lower/faster/higher for P compared with TR, 3--identical for P and TR, 4--slightly lower/faster/higher for TR compared with P, and 5--markedly lower/faster/higher for TR compared with P.Liver segments II/III, II-IV and V-VIII served in 6, 3, and 7 donors as transplanted liver segments. Volumes were 642.9±368.8 ml for TR with vessels, 623.8±349.1 ml for P with vessels, and 605.2±345.8 ml for P without vessels (P<0.01. Regression equations between intraoperative weights and volumes were y = 0.94x+30.1 (R2 = 0.92; P<0.001 for TR with vessels, y = 1.00x+12.0 (R2 = 0.92; P<0.001 for P with vessels, and y = 1.01x+28.0 (R2 = 0.92; P<0.001 for P without vessels. Inter-observer agreement showed a bias of 1.8 ml for TR with vessels, 5.4 ml for P with vessels, and 4.6 ml for P without vessels. For the degree of manual correction, speed for completion and overall intuitiveness, scale values were 2.6±0.8, 2.4±0.5 and 2.CT-volumetry performed with P can predict accurately graft

  15. Accuracy of automated volumetry of pulmonary nodules across different multislice CT scanners

    International Nuclear Information System (INIS)

    Das, Marco; Muehlenbruch, Georg; Mahnken, Andreas H.; Katoh, Markus; Guenther, Rolf W.; Wildberger, Joachim E.; Ley-Zaporozhan, Julia; Kauczor, Hans-Ulrich; Gietema, H.A.; Prokop, Mathias; Czech, Andre; Diederich, Stefan; Bakai, Annemarie; Salganicoff, Marcos

    2007-01-01

    The purpose of this study was to compare the accuracy of an automated volumetry software for phantom pulmonary nodules across various 16-slice multislice spiral CT (MSCT) scanners from different vendors. A lung phantom containing five different nodule categories (intraparenchymal, around a vessel, vessel attached, pleural, and attached to the pleura), with each category comprised of 7-9 nodules (total, n = 40) of varying sizes (diameter 3-10 mm; volume 6.62 mm 3 -525 mm 3 ), was scanned with four different 16-slice MSCT scanners (Siemens, GE, Philips, Toshiba). Routine and low-dose chest protocols with thin and thick collimations were applied. The data from all scanners were used for further analysis using a dedicated prototype volumetry software. Absolute percentage volume errors (APE) were calculated and compared. The mean APE for all nodules was 8.4% (±7.7%) for data acquired with the 16-slice Siemens scanner, 14.3% (±11.1%) for the GE scanner, 9.7% (±9.6%) for the Philips scanner and 7.5% (±7.2%) for the Toshiba scanner, respectively. The lowest APEs were found within the diameter size range of 5-10 mm and volumes >66 mm 3 . Nodule volumetry is accurate with a reasonable volume error in data from different scanner vendors. This may have an important impact for intraindividual follow-up studies. (orig.)

  16. Requirements on the Redshift Accuracy for future Supernova and Number Count Surveys

    International Nuclear Information System (INIS)

    Huterer, Dragan; Kim, Alex; Broderick, Tamara

    2004-01-01

    We investigate the required redshift accuracy of type Ia supernova and cluster number-count surveys in order for the redshift uncertainties not to contribute appreciably to the dark energy parameter error budget. For the SNAP supernova experiment, we find that, without the assistance of ground-based measurements, individual supernova redshifts would need to be determined to about 0.002 or better, which is a challenging but feasible requirement for a low-resolution spectrograph. However, we find that accurate redshifts for z < 0.1 supernovae, obtained with ground-based experiments, are sufficient to immunize the results against even relatively large redshift errors at high z. For the future cluster number-count surveys such as the South Pole Telescope, Planck or DUET, we find that the purely statistical error in photometric redshift is less important, and that the irreducible, systematic bias in redshift drives the requirements. The redshift bias will have to be kept below 0.001-0.005 per redshift bin (which is determined by the filter set), depending on the sky coverage and details of the definition of the minimal mass of the survey. Furthermore, we find that X-ray surveys have a more stringent required redshift accuracy than Sunyaev-Zeldovich (SZ) effect surveys since they use a shorter lever arm in redshift; conversely, SZ surveys benefit from their high redshift reach only so long as some redshift information is available for distant (zgtrsim1) clusters

  17. Verification of a CT scanner using a miniature step gauge

    DEFF Research Database (Denmark)

    Cantatore, Angela; Andreasen, J.L.; Carmignato, S.

    2011-01-01

    The work deals with performance verification of a CT scanner using a 42mm miniature replica step gauge developed for optical scanner verification. Errors quantification and optimization of CT system set-up in terms of resolution and measurement accuracy are fundamental for use of CT scanning...

  18. Applications of SPECT / CT at the Institute of Nephrology

    International Nuclear Information System (INIS)

    Zayas Crespo, Francisco; Batlle Leal, Fernando; Mesa Dueñas, Niurka

    2016-01-01

    The main applications of SPECT-CT in the diagnosis of various diseases, as well as the experiences gained from the operation of this hybrid system are described. To date we have studied more than 100 patients from different hospitals in the capital and other nearby provinces, including. Including major diseases: kidney, bone, oncological and endocrine. So far, the results achieved have enabled diagnostic evaluation of higher accuracy and detail, even improving in a significant number of patients the initial diagnosis or modifying this fully. (author)

  19. Detection ability of FDG-PET/CT comparing with other imaging modalities in multiple myeloma patients

    International Nuclear Information System (INIS)

    Chae, Min Jeong; Lee, Tae Hyun; Pai, Moon Sun; Cheon, Gi Jeong; Choi, Chang Woon; Lim, Sang Moo

    2007-01-01

    Multiple myeloma (MM) is characterized by bone marrow infiltration with malignant plasma cells. It is important to detect involving bone for diagnosis and management of MM. The aim of this study was to evaluate the diagnostic ability and limitation of 18F-FDG-PET/CT (PET/CT) comparing other imaging modalities (separated PET and CT, whole body plain X-ray (XR), bone scintigraphy (BS), and MRI) in MM. Twenty PET/CT scans were performed in 16 patients (M: F=6: 10, median age=59 y). PET/CT findings were compared with available other images (n of CT=21, XR=21, BS=8, and MRI=5). Concordance with more than 2 image modalities, laboratory data, symptom, and biopsies were used for diagnosis of detected lesions. PET/CT revealed 256 of total 287 sites (sensitivity, 89.2%; accuracy, 84.8%). The sensitivity and accuracy of separating PET, CT, and XR were 86.3%, 70.4%; 47.4%, 50.3%; and 72.8%, 72.4%, respectively. Available BS identified 67 of 202 sites (sensitivity, 33.2%; accuracy, 44.0%). MRI detected 20 of 24 sites (sensitivity, 83.3%; accuracy, 36.3%). False positive rate (FP) of PET, XR, and MRI was as high as 87.8%, 95.1%, and 100%. PET for rib lesion identified 9 of 10 patients (90.0%) but for skull lesion only 4 of 7 patients (57.2%) with underestimation. 5 patients in MRI showed diffuse marrow signal change but only 3 had marrow involvement. But PET/CT showed higher accuracy than MRI. PET/CT was the most useful tool for detecting involving bone of MM comparing with other imaging modalities. Moreover, PET/CT is expected to overcome the limitations for the small osteolytic bone lesions with diffuse FDG uptake on PET

  20. Adenocarcinoma of the pancreatic head: preoperative helical CT. Criteria of resectability

    International Nuclear Information System (INIS)

    Kozima, Shigeru; Szelagowski, Carlos; Tisserand, Guy L.; Ocampo, Carlos; Zandalazini, Hugo; Silva, Walter; Oria, Alejandro; Vidovic, Gustavo; Varas, Pablo

    2001-01-01

    Objective: The purpose of this study is to determine the accuracy of biphasic helical CT scanning in predicting resectability of adenocarcinoma of the head of the pancreas by staying tumor involvement of the portal and superior mesenteric veins. Material and methods: 46 patients with proven adenocarcinoma of the head of the pancreas who underwent curative or palliative surgery were studied with preoperative biphasic helical CT scanning. Tumor involvement of the portal and mesenteric veins was graduated on a 1-3 scale based on circumferential contiguity of the tumor vessel. Grade 1: without contact; grade 2: tumor involvement of less than 50% of the vessel; grade 3: tumor involvement of more than 50%. Results: The total number of vessels evaluated was 92. In our series the preoperative biphasic helical CT was accurate in 77% for resectability and unresectability. Conclusion: Our experience of staging in 3 grades with biphasic helical CT, vessel involvement the portal and superior mesenteric veins of adenocarcinoma of the head of the pancreas is highly specific for unresectable tumor in patients who were graded 2 and 3. (author)

  1. CT of abdominal abscesses

    International Nuclear Information System (INIS)

    Korobkin, M.T.

    1987-01-01

    The imaging search for a suspected abdominal abscess is common in hospitalized patients, especially after recent abdominal surgery. This paper examines the role of CT in the detection, localization, and treatment of abdominal abscess. The accuracy, limitations, and technical aspects of CT in this clinical setting are discussed. The diagnosis of an abscess is based on the demonstration of a circumscribed abnormal fluid collection. Although percutaneous aspiration with gram stain and culture is usually indicated to differentiate abscess from other fluid collections, the CT-based detection of extraluminal gas bubbles makes the diagnosis of an abscess highly likely. CT is compared with conventional radiographic studies, US, and radio-nuclide imaging. Specific CT and clinical features of abscesses in the following sites are emphasized: subphrenic space, liver, pancreas, kidneys, psoas muscle, appendix, and colonic diverticula. Most abdominal abscesses can be successfully treated with percutaneous drainage techniques. The techniques, results, and limitations of percutaneous abscess drainage are reviewed

  2. T-staging of gastric cancer of air-filling multidetector-row CT: Comparison with hydro-multidetector-row CT

    Energy Technology Data Exchange (ETDEWEB)

    Kumano, Seishi, E-mail: kumano@radiol.med.kindai.ac.jp [Department of Radiology, Kinki University School of Medicine, Osakasayama, Osaka 589-8511 (Japan); Okada, Masahiro; Shimono, Taro; Kuwabara, Masatomo; Yagyu, Yukinobu; Imaoka, Izumi; Ashikaga, Ryuichiro; Ishii, Kazunari; Murakami, Takamichi [Department of Radiology, Kinki University School of Medicine, Osakasayama, Osaka 589-8511 (Japan)

    2012-11-15

    Purpose: The purpose of this study was to evaluate the accuracy of T-staging of gastric cancer by air-filling multidetector-row CT (air-MDCT) compared with water-filling MDCT (hydro-MDCT). Materials and methods: One hundred fifteen patients with histologically diagnosed gastric cancer were included in this study. Fifty-eight patients underwent air-MDCT, and the remaining 57 had hydro-MDCT using a 64-channel scanner. Based on the volumetric data of contrast-enhanced MDCT obtained about 75 s after intravenously injecting 525 mg iodine per kilogram patients weight (525 mgI/kg) nonionic contrast material at the rate of 2 ml/s, oblique coronal and oblique sagittal multi-planar reformatted images perpendicular to the stomach wall, including the tumor, were reconstructed on a workstation. Mural invasion of gastric cancer into the gastric wall, as visualized by CT, was classified according to the TNM classification, and the results of T-staging by MDCT were compared with those by pathologic analysis after surgery. Results: Correct assessment of T-staging by air-CT was achieved in 48 of 58 patients (83%), and that by hydro-MDCT was 49 of 57 patients (86%). The sensitivity, specificity, and accuracy of the technique in determining the invasion of serosa were 88%, 93%, and 91% for air-CT and 83%, 95%, and 91% for hydro-CT. There were no significant differences between hydro-MDCT and air-MDCT in sensitivity (P = 0.73), specificity (P = 0.71) and accuracy (P = 0.98). Conclusion: Air-MDCT is a very valuable tool in T-staging of gastric cancer as well as hydro-MDCT.

  3. Graphics Processing Unit-Accelerated Nonrigid Registration of MR Images to CT Images During CT-Guided Percutaneous Liver Tumor Ablations.

    Science.gov (United States)

    Tokuda, Junichi; Plishker, William; Torabi, Meysam; Olubiyi, Olutayo I; Zaki, George; Tatli, Servet; Silverman, Stuart G; Shekher, Raj; Hata, Nobuhiko

    2015-06-01

    Accuracy and speed are essential for the intraprocedural nonrigid magnetic resonance (MR) to computed tomography (CT) image registration in the assessment of tumor margins during CT-guided liver tumor ablations. Although both accuracy and speed can be improved by limiting the registration to a region of interest (ROI), manual contouring of the ROI prolongs the registration process substantially. To achieve accurate and fast registration without the use of an ROI, we combined a nonrigid registration technique on the basis of volume subdivision with hardware acceleration using a graphics processing unit (GPU). We compared the registration accuracy and processing time of GPU-accelerated volume subdivision-based nonrigid registration technique to the conventional nonrigid B-spline registration technique. Fourteen image data sets of preprocedural MR and intraprocedural CT images for percutaneous CT-guided liver tumor ablations were obtained. Each set of images was registered using the GPU-accelerated volume subdivision technique and the B-spline technique. Manual contouring of ROI was used only for the B-spline technique. Registration accuracies (Dice similarity coefficient [DSC] and 95% Hausdorff distance [HD]) and total processing time including contouring of ROIs and computation were compared using a paired Student t test. Accuracies of the GPU-accelerated registrations and B-spline registrations, respectively, were 88.3 ± 3.7% versus 89.3 ± 4.9% (P = .41) for DSC and 13.1 ± 5.2 versus 11.4 ± 6.3 mm (P = .15) for HD. Total processing time of the GPU-accelerated registration and B-spline registration techniques was 88 ± 14 versus 557 ± 116 seconds (P processing time. The GPU-accelerated volume subdivision technique may enable the implementation of nonrigid registration into routine clinical practice. Copyright © 2015 AUR. Published by Elsevier Inc. All rights reserved.

  4. Comparison of diagnostic accuracy of {sup 111}In-pentetreotide SPECT and {sup 68}Ga-DOTATOC PET/CT: A lesion-by-lesion analysis in patients with metastatic neuroendocrine tumours

    Energy Technology Data Exchange (ETDEWEB)

    Binnebeek, S. van; Vanbilloen, B.; Baete, K.; Terwinghe, C.; Koole, M.; Mortelmans, L. [KU Leuven, Nuclear Medicine, University Hospitals Leuven and Department of Imaging and Pathology, Leuven (Belgium); Mottaghy, F.M. [Maastricht University Medical Center, Department of Nuclear Medicine, Maastricht (Netherlands); University Hospital Aachen, Department of Nuclear Medicine, Aachen (Germany); Clement, P.M. [KU Leuven, Medical Oncology, University Hospitals Leuven and Laboratory of Experimental Oncology, Leuven (Belgium); Bogaerts, K. [KU Leuven and UHasselt, Department of Public Health and Primary Care (I-BioStat), Leuven (Belgium); Haustermans, K. [KU Leuven, Radiation Oncology, University Hospitals Leuven and Department of Oncology, Leuven (Belgium); Nackaerts, K. [University Hospitals Leuven, Pulmonology, Leuven (Belgium); Cutsem, E. van; Verslype, C. [KU Leuven, Division of Digestive Oncology, University Hospitals Leuven and Department of Oncology, Leuven (Belgium); Verbruggen, A. [KU Leuven, Laboratory for Radiopharmacy, Leuven (Belgium); Deroose, C.M. [KU Leuven, Nuclear Medicine, University Hospitals Leuven and Department of Imaging and Pathology, Leuven (Belgium); University Hospitals Leuven, Nuclear Medicine, Leuven (Belgium)

    2016-03-15

    To compare the diagnostic accuracy of {sup 111}In-pentetreotide-scintigraphy with {sup 68}Ga-DOTATOC-positron emission tomography (PET)/computed tomography (CT) in patients with metastatic-neuroendocrine tumour (NET) scheduled for peptide receptor radionuclide therapy (PRRT). Incremental lesions (ILs) were defined as lesions observed on only one modality. Fifty-three metastatic-NET-patients underwent {sup 111}In-pentetreotide-scintigraphy (24 h post-injection; planar+single-photon emission CT (SPECT) abdomen) and whole-body {sup 68}Ga-DOTATOC-PET/CT. SPECT and PET were compared in a lesion-by-lesion and organ-by-organ analysis, determining the total lesions and ILs for both modalities. Significantly more lesions were detected on {sup 68}Ga-DOTATOC-PET/CT versus {sup 111}In-pentetreotide-scintigraphy. More specifically, we observed 1,098 lesions on PET/CT (range: 1-105; median: 15) versus 660 on SPECT (range: 0-73, median: 9) (p<0.0001), with 439 PET-ILs (42/53 patients) and one SPECT-IL (1/53 patients). The sensitivity for PET/CT was 99.9 % (95 % CI, 99.3-100.0), for SPECT 60.0 % (95 % CI, 48.5-70.2). The organ-by-organ analysis showed that the PET-ILs were most frequently visualized in liver and skeleton. Ga-DOTATOC-PET/CT is superior for the detection of NET-metastases compared to {sup 111}In-pentetreotide SPECT. (orig.)

  5. Anatomical accuracy of lesion localization. Retrospective interactive rigid image registration between 18F-FDG-PET and X-ray CT

    International Nuclear Information System (INIS)

    Noemayr, A.; Roemer, W.; Kuwert, T.; Hothorn, T.; Pfahlberg, A.; Hornegger, J.; Bautz, W.

    2005-01-01

    The aim of this study was to evaluate the anatomical accuracy and reproducibility of retrospective interactive rigid image registration (RIR) between routinely archived X-ray computer tomography (CT) and positron emission tomography performed with 18 F-deoxyglucose (FDG-PET) in oncological patients. Methods: two observers registered PET and CT data obtained in 37 patients using a commercially available image fusion tool. RIR was performed separately for the thorax and the abdomen using physiological FDG uptake in several organs as a reference. One observer performed the procedure twice (01a and 01b), another person once (02). For 94 malignant lesions, clearly visible in CT and PET, the signed and absolute distances between their representation on PET and CT were measured in X-, Y-, and Z-direction with reference to a coordinate system centered in the CT representation of each lesion (X-, Y-, Z-distances). Results: the mean differences of the signed and absolute distances between 01a, 01b, and 02 did not exceed 3 mm in any dimension. The absolute X-, Y-, and Z-distances ranged between 0.57 ± 0.58 cm for 01a (X-direction) and 1.12 ± 1.28 cm for 02 (Z-direction). When averaging the absolute distances measured by 01a, 01b, and 02, the percentage of lesions misregistered by less than 1.5 cm was 91% for the X-, 88% for the Y-, and 77% for the Z-direction. The larger error of fusion determined for the remaining lesions was caused by non-rigid body transformations due to differences in breathing, arm position, or bowel movements between the two examinations. Mixed effects analysis of the signed and absolute X-, Y-, and Z-distances disclosed a significantly greater misalignment in the thorax than in the abdomen as well as axially than transaxially. Conclusion: the anatomical inaccuracy of RIR can be expected to be <1.5 cm for the majority of neoplastic foci. errors of alignment are bigger in the thorax and in Z-direction, due to non-rigid body transformations caused, e

  6. Detection and compensation of organ/lesion motion using 4D-PET/CT respiratory gated acquisition techniques

    International Nuclear Information System (INIS)

    Bettinardi, Valentino; Picchio, Maria; Di Muzio, Nadia; Gianolli, Luigi; Gilardi, Maria Carla; Messa, Cristina

    2010-01-01

    Purpose: To describe the degradation effects produced by respiratory organ and lesion motion on PET/CT images and to define the role of respiratory gated (RG) 4D-PET/CT techniques to compensate for such effects. Methods: Based on the literature and on our own experience, technical recommendations and clinical indications for the use of RG 4D PET/CT have been outlined. Results: RG 4D-PET/CT techniques require a state of the art PET/CT scanner, a respiratory monitoring system and dedicated acquisition and processing protocols. Patient training is particularly important to obtain a regular breathing pattern. An adequate number of phases has to be selected to balance motion compensation and statistical noise. RG 4D PET/CT motion free images may be clinically useful for tumour tissue characterization, monitoring patient treatment and target definition in radiation therapy planning. Conclusions: RG 4D PET/CT is a valuable tool to improve image quality and quantitative accuracy and to assess and measure organ and lesion motion for radiotherapy planning.

  7. Improving Accuracy for Image Fusion in Abdominal Ultrasonography

    Directory of Open Access Journals (Sweden)

    Caroline Ewertsen

    2012-08-01

    Full Text Available Image fusion involving real-time ultrasound (US is a technique where previously recorded computed tomography (CT or magnetic resonance images (MRI are reformatted in a projection to fit the real-time US images after an initial co-registration. The co-registration aligns the images by means of common planes or points. We evaluated the accuracy of the alignment when varying parameters as patient position, respiratory phase and distance from the co-registration points/planes. We performed a total of 80 co-registrations and obtained the highest accuracy when the respiratory phase for the co-registration procedure was the same as when the CT or MRI was obtained. Furthermore, choosing co-registration points/planes close to the area of interest also improved the accuracy. With all settings optimized a mean error of 3.2 mm was obtained. We conclude that image fusion involving real-time US is an accurate method for abdominal examinations and that the accuracy is influenced by various adjustable factors that should be kept in mind.

  8. Brain tumor and CT, 1. Relationship between the consistency of a brain tumor and the CT findings

    Energy Technology Data Exchange (ETDEWEB)

    Suzuki, N.; Katada, K.; Shinomiya, Y.; Sano, H.; Kanno, T. (Fujita Gakuen Univ., School of Medicine, Toyoake, Aichi (Japan))

    1981-08-01

    It is very important for a neurosurgeon to know the consistency of a brain tumor preoperatively, since the information is of much use in indicating the likely difficulty of the operation, which operative tools should be selected, the amount of bleeding to be expected from the tumor, and so on. The authors, therefore, tried to evaluate the consistency of brain tumors preoperatively. Twenty-seven cases in which the margin of the tumor was made clear with a homogeneous stain were studied concerning the relationship between the tumor consistency and the CT findings. The results are as follows: 1) A higher CT number on a plain CT indicated a harder consistency of the tumor. 2) A lesser contrast index (CT number on enhancement CT/CT number on plain CT) showed a harder consistency of the tumor.

  9. Sensitivity and accuracy of volumetry of pulmonary nodules on low-dose 16- and 64-row multi-detector CT: an anthropomorphic phantom study

    International Nuclear Information System (INIS)

    Xie, Xueqian; Zhao, Yingru; Ooijen, Peter M.A. van; Vliegenthart, Rozemarijn; Snijder, Roland A.; Greuter, Marcel J.W.; Jong, Pim A. de; Oudkerk, Matthijs; Bock, Geertruida H. de

    2013-01-01

    To assess the sensitivity of detection and accuracy of volumetry by manual and semi-automated quantification of artificial pulmonary nodules in an anthropomorphic thoracic phantom on low-dose CT. Fifteen artificial spherical nodules (diameter 3, 5, 8, 10 and 12 mm; CT densities -800, -630 and +100 HU) were randomly placed inside an anthropomorphic thoracic phantom. The phantom was examined on 16- and 64-row multidetector CT with a low-dose protocol. Two independent blinded observers screened for pulmonary nodules. Nodule diameter was measured manually, and volume calculated. For solid nodules (+100 HU), diameter and volume were also evaluated by semi-automated software. Differences in observed volumes between the manual and semi-automated method were evaluated by a t-test. Sensitivity was 100 % for all nodules of >5 mm and larger, 60-80 % for solid and 0-20 % for non-solid 3-mm nodules. No false-positive nodules but high inter-observer reliability and inter-technique correlation were found. Volume was underestimated manually by 24.1 ± 14.0 % for nodules of any density, and 26.4 ± 15.5 % for solid nodules, compared with 7.6 ± 8.5 % (P 5 mm in diameter. Semi-automated volumetry yielded more accurate nodule volumes than manual measurements. (orig.)

  10. A phantom for testing of 4D-CT for radiotherapy of small lesions

    International Nuclear Information System (INIS)

    Dunn, L.; Kron, T.; Taylor, M. L.; Callahan, J.; Franich, R. D.

    2012-01-01

    Purpose: The use of time-resolved four-dimensional computed tomography (4D-CT) in radiotherapy requires strict quality assurance to ensure the accuracy of motion management protocols. The aim of this work was to design and test a phantom capable of large amplitude motion for use in 4D-CT, with particular interest in small lesions typical for stereotactic body radiotherapy. Methods: The phantom of “see-saw” design is light weight, capable of including various sample materials and compatible with several surrogate marker signal acquisition systems. It is constructed of polymethylmethacrylate (Perspex) and its movement is controlled via a dc motor and drive wheel. It was tested using two CT scanners with different 4D acquisition methods: the Philips Brilliance Big Bore CT (helical scan, pressure belt) and a General Electric Discovery STE PET/CT (axial scan, infrared marker). Amplitudes ranging from 1.5 to 6.0 cm and frequencies of up to 40 cycles per minute were used to study the effect of motion on image quality. Maximum intensity projections (MIPs), as well as average intensity projections (AIPs) of moving objects were investigated and their quality dependence on the number of phase reconstruction bins assessed. Results: CT number discrepancies between moving and stationary objects were found to have no systematic dependence on amplitude, frequency, or specific interphase variability. MIP-delineated amplitudes of motion were found to match physical phantom amplitudes to within 2 mm for all motion scenarios tested. Objects undergoing large amplitude motions (>3.0 cm) were shown to cause artefacts in MIP and AIP projections when ten phase bins were assigned. This problem can be mitigated by increasing the number of phase bins in a 4D-CT scan. Conclusions: The phantom was found to be a suitable tool for evaluating the image quality of 4D-CT motion management technology, as well as providing a quality assurance tool for intercenter/intervendor testing of commercial

  11. SU-E-T-70: Commissioning a Multislice CT Scanner for X-Ray CT Polymer Gel Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Johnston, H [University of Victoria, Victoria, BC (Australia); UT Southwestern Medical Center, Dallas, TX (United States); Hilts, M [University of Victoria, Victoria, BC (Australia); BC Cancer Agency, Kelowna, BC (Australia); Jirasek, A [University of Victoria, Victoria, BC (Australia)

    2014-06-01

    Purpose: To commission a multislice computed tomography (CT) scanner for fast and reliable readout of radiation therapy (RT) dose distributions using CT polymer gel dosimetry (PGD). Methods: Commissioning was performed for a 16-slice CT scanner using images acquired through a 1L cylinder filled with water. Additional images were collected using a single slice machine for comparison purposes. The variability in CT number associated with the anode heel effect was evaluated and used to define a new slice-by-slice background image subtraction technique. Image quality was assessed for the multislice system by comparing image noise and uniformity to that of the single slice machine. The consistency in CT number across slices acquired simultaneously using the multislice detector array was also evaluated. Finally, the variability in CT number due to increasing x-ray tube load was measured for the multislice scanner and compared to the tube load effects observed on the single slice machine. Results: Slice-by-slice background subtraction effectively removes the variability in CT number across images acquired simultaneously using the multislice scanner and is the recommended background subtraction method when using a multislice CT system. Image quality for the multislice machine was found to be comparable to that of the single slice scanner. Further study showed CT number was consistent across image slices acquired simultaneously using the multislice detector array for each detector configuration of the slice thickness examined. In addition, the multislice system was found to eliminate variations in CT number due to increasing x-ray tube load and reduce scanning time by a factor of 4 when compared to imaging a large volume using a single slice scanner. Conclusion: A multislice CT scanner has been commissioning for CT PGD, allowing images of an entire dose distribution to be acquired in a matter of minutes. Funding support provided by the Natural Sciences and Engineering

  12. 68Ga-DOTA-NOC PET/CT in comparison with CT for the detection of bone metastasis in patients with neuroendocrine tumours

    International Nuclear Information System (INIS)

    Ambrosini, Valentina; Nanni, Cristina; Castellucci, Paolo; Allegri, Vincenzo; Montini, Giancarlo; Franchi, Roberto; Zompatori, Maurizio; Campana, Davide; Tomassetti, Paola; Rubello, Domenico; Fanti, Stefano

    2010-01-01

    To retrospectively evaluate the sensitivity, specificity and accuracy of 68 Ga-DOTA-NOC PET/CT and CT alone for the evaluation of bone metastasis in patients with neuroendocrine tumour (NET). From among patients with NET who underwent 68 Ga-DOTA-NOC PET/CT between April 2006 and November 2008 in our centre, 223 were included in the study. Criteria for inclusion were pathological confirmation of NET and a follow-up period of at least 10 months. PET and CT images were retrospectively reviewed by two nuclear medicine specialists and two radiologists, respectively, without knowledge of the patient history or the findings of other imaging modalities. PET data were compared with the CT findings. Interobserver agreement was evaluated in terms of the kappa score. Clinical and imaging follow-up were used as the standard of reference to evaluate the PET findings. PET was performed for staging (49/223), unknown primary tumour detection (24/223), restaging (32/223), restaging before radioimmunotherapy (1/223), evaluation during therapy (12/223), equivocal findings on conventional imaging (4/223 at the bone level; 61/223 at sites other than bone), and follow-up (40/223). A very high interobserver agreement was observed. CT detected at least one bone lesion in only 35 of 44 patients with a positive PET scan. In particular, PET showed more lesions in 20/35 patients, a lower number of lesions in 8/35, and the same number in 7/35. The characteristics of the lesions (sclerotic, lytic, mixed) on the basis of the CT report did not influence PET reading. PET revealed the presence of at least one bone metastasis in nine patients with a negative CT scan. Considering patients with a negative PET scan (179), CT showed equivocal findings at the bone level in three (single small sclerotic abnormality in two at the spine level, and bilateral small sclerotic abnormalities in the humeri, femurs and scapula). Clinical follow-up confirmed the PET findings in all patients; thus there were no false

  13. Clinical significance of CT in the preoperative diagnosis of the staging of rectal cancer patients

    International Nuclear Information System (INIS)

    Itano, Satoshi; Fuchimoto, Sadanori; Hamada, Fumihiro; Kimura, Takanobu; Iwagaki, Hiromi; Maeda, Tetsuya; Orita, Kunzo

    1987-01-01

    The value of computed tomography (CT) in the preoperative clinical staging of rectal cancer was prospectively studied in 28 patients with macroscopically proven cancer. The CT studies were based on the previously established CT diagnostic criteria for wall invasion (S factor), lymph node metastases (N factor), and liver metastases (H factor). When macroscopic findings were used as the standard, the accuracy of CT was 61 % for S factor, 32 % for N factor, and 86 % for H factor. Using histological findings as the standard, the accuracy was 48 % for S factor and 16 % for N factor. Overall, CT had a high accuracy for H factor in all sites of cancer (75 % in the rectosigmoid, 86 % in the area above the peritoneal reflection, and 90 % in the area below the peritoneal reflection). For the other S and N factors, CT seemed to be of limited value in the preoperative diagnosis. (Namekawa, K.)

  14. Iterative metal artifact reduction improves dose calculation accuracy. Phantom study with dental implants

    Energy Technology Data Exchange (ETDEWEB)

    Maerz, Manuel; Mittermair, Pia; Koelbl, Oliver; Dobler, Barbara [Regensburg University Medical Center, Department of Radiotherapy, Regensburg (Germany); Krauss, Andreas [Siemens Healthcare GmbH, Forchheim (Germany)

    2016-06-15

    Metallic dental implants cause severe streaking artifacts in computed tomography (CT) data, which affect the accuracy of dose calculations in radiation therapy. The aim of this study was to investigate the benefit of the metal artifact reduction algorithm iterative metal artifact reduction (iMAR) in terms of correct representation of Hounsfield units (HU) and dose calculation accuracy. Heterogeneous phantoms consisting of different types of tissue equivalent material surrounding metallic dental implants were designed. Artifact-containing CT data of the phantoms were corrected using iMAR. Corrected and uncorrected CT data were compared to synthetic CT data to evaluate accuracy of HU reproduction. Intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) plans were calculated in Oncentra v4.3 on corrected and uncorrected CT data and compared to Gafchromic trademark EBT3 films to assess accuracy of dose calculation. The use of iMAR increased the accuracy of HU reproduction. The average deviation of HU decreased from 1006 HU to 408 HU in areas including metal and from 283 HU to 33 HU in tissue areas excluding metal. Dose calculation accuracy could be significantly improved for all phantoms and plans: The mean passing rate for gamma evaluation with 3 % dose tolerance and 3 mm distance to agreement increased from 90.6 % to 96.2 % if artifacts were corrected by iMAR. The application of iMAR allows metal artifacts to be removed to a great extent which leads to a significant increase in dose calculation accuracy. (orig.) [German] Metallische Implantate verursachen streifenfoermige Artefakte in CT-Bildern, welche die Dosisberechnung beeinflussen. In dieser Studie soll der Nutzen des iterativen Metall-Artefakt-Reduktions-Algorithmus iMAR hinsichtlich der Wiedergabetreue von Hounsfield-Werten (HU) und der Genauigkeit von Dosisberechnungen untersucht werden. Es wurden heterogene Phantome aus verschiedenen Arten gewebeaequivalenten Materials mit

  15. Parametric boundary reconstruction algorithm for industrial CT metrology application.

    Science.gov (United States)

    Yin, Zhye; Khare, Kedar; De Man, Bruno

    2009-01-01

    High-energy X-ray computed tomography (CT) systems have been recently used to produce high-resolution images in various nondestructive testing and evaluation (NDT/NDE) applications. The accuracy of the dimensional information extracted from CT images is rapidly approaching the accuracy achieved with a coordinate measuring machine (CMM), the conventional approach to acquire the metrology information directly. On the other hand, CT systems generate the sinogram which is transformed mathematically to the pixel-based images. The dimensional information of the scanned object is extracted later by performing edge detection on reconstructed CT images. The dimensional accuracy of this approach is limited by the grid size of the pixel-based representation of CT images since the edge detection is performed on the pixel grid. Moreover, reconstructed CT images usually display various artifacts due to the underlying physical process and resulting object boundaries from the edge detection fail to represent the true boundaries of the scanned object. In this paper, a novel algorithm to reconstruct the boundaries of an object with uniform material composition and uniform density is presented. There are three major benefits in the proposed approach. First, since the boundary parameters are reconstructed instead of image pixels, the complexity of the reconstruction algorithm is significantly reduced. The iterative approach, which can be computationally intensive, will be practical with the parametric boundary reconstruction. Second, the object of interest in metrology can be represented more directly and accurately by the boundary parameters instead of the image pixels. By eliminating the extra edge detection step, the overall dimensional accuracy and process time can be improved. Third, since the parametric reconstruction approach shares the boundary representation with other conventional metrology modalities such as CMM, boundary information from other modalities can be directly

  16. CT findings of tuberculous lymphadenifis in parotid gland

    International Nuclear Information System (INIS)

    Wang Changfu; Wang Binjie; Zhang Heping; Jin Haiying; Nie Peng; Chang Liang; Wei Haigang; Zou Ling

    2008-01-01

    Objective: To analyze the CT findings of tuberculous lymphadenitis in parotid gland, so as to improve the diagnostic accuracy of tuberculosis of parotid gland. Methods: Nine cases with tuberculous lymphadenitis in parotid gland confirmed by surgical pathology and acid-fast bacilli after preoperative spiral CT plain scan and two phases dynamic enhancement scan were retrospectively analyzed. Imaging findings of CT were reviewed and compared with surgical pathology. Results: Seven of the 9 cases of tuberculosis of the parotid gland occurred in the left side, and 2 in the right side, and superficial lobe involvement occurred in 8 cases and deep lobe in 1 case. The lesion was classified as tumour type (8 cases) and infiltration type (1 case). In turnout type, the number of lesion was from 1 to 4, and the size was from 2.7 to 5.3 cm in diameter. One case of infiltration type measured 3.4 cm in diameter. On CT plain scan, the lesions showed homogeneous slight high-density with regular edge in 5 cases and irregular low-density in 4 cases, and 2 of them with partly blurred edge. On CT enhanced scan, uniform moderate enhancement was seen in 3 cases, circular enhancement in 4 cases, inhomogeneous enhancement in 1 case, and lace-like enhancement in 1 case. Local infiltration occurred in 6 cases. Lymphadenovarix in the same side of lesion occurred in 2 cases. Conclusion: CT findings of tuberculous lymphadenitis in parotid gland present diversification, which correlate well with pathological changes. Understanding of characteristic CT findings of tuberculous lymphadenitis in parotid gland is helpful for differential diagnosis, but final diagnosis still depends on pathology and acid-fast bacilli. (authors)

  17. Non-rigid CT/CBCT to CBCT registration for online external beam radiotherapy guidance

    Science.gov (United States)

    Zachiu, Cornel; de Senneville, Baudouin Denis; Tijssen, Rob H. N.; Kotte, Alexis N. T. J.; Houweling, Antonetta C.; Kerkmeijer, Linda G. W.; Lagendijk, Jan J. W.; Moonen, Chrit T. W.; Ries, Mario

    2018-01-01

    Image-guided external beam radiotherapy (EBRT) allows radiation dose deposition with a high degree of accuracy and precision. Guidance is usually achieved by estimating the displacements, via image registration, between cone beam computed tomography (CBCT) and computed tomography (CT) images acquired at different stages of the therapy. The resulting displacements are then used to reposition the patient such that the location of the tumor at the time of treatment matches its position during planning. Moreover, ongoing research aims to use CBCT-CT image registration for online plan adaptation. However, CBCT images are usually acquired using a small number of x-ray projections and/or low beam intensities. This often leads to the images being subject to low contrast, low signal-to-noise ratio and artifacts, which ends-up hampering the image registration process. Previous studies addressed this by integrating additional image processing steps into the registration procedure. However, these steps are usually designed for particular image acquisition schemes, therefore limiting their use on a case-by-case basis. In the current study we address CT to CBCT and CBCT to CBCT registration by the means of the recently proposed EVolution registration algorithm. Contrary to previous approaches, EVolution does not require the integration of additional image processing steps in the registration scheme. Moreover, the algorithm requires a low number of input parameters, is easily parallelizable and provides an elastic deformation on a point-by-point basis. Results have shown that relative to a pure CT-based registration, the intrinsic artifacts present in typical CBCT images only have a sub-millimeter impact on the accuracy and precision of the estimated deformation. In addition, the algorithm has low computational requirements, which are compatible with online image-based guidance of EBRT treatments.

  18. CT, MRI and PET image fusion using the ProSoma 3D simulation software

    International Nuclear Information System (INIS)

    Dalah, E.; Bradley, D.A.; Nisbet, A.; Reise, S.

    2008-01-01

    Full text: Multi-modality imaging is involved in almost all oncology applications focusing on the extent of disease and target volume delineation. Commercial image fusion software packages are becoming available but require comprehensive evaluation to ensure reliability of fusion and the underpinning registration algorithm particularly for radiotherapy. The present work seeks to assess such accuracy for a number of available registration methods provided by the commercial package ProSoma. A NEMA body phantom was used in evaluating CT, MR and PET images. In addition, discussion is provided concerning the choice and geometry of fiducial markers in phantom studies and the effect of window-level on target size, in particular in regard to the application of multi modality imaging in treatment planning. In general, the accuracy of fusion of multi-modality images was within 0.5-1.5 mm of actual feature diameters and < 2 ml volume of actual values, particularly in CT images. (author)

  19. SU-F-I-08: CT Image Ring Artifact Reduction Based On Prior Image

    Energy Technology Data Exchange (ETDEWEB)

    Yuan, C; Qi, H; Chen, Z; Wu, S; Xu, Y; Zhou, L [Southern Medical University, Guangzhou, Guangdong (China)

    2016-06-15

    Purpose: In computed tomography (CT) system, CT images with ring artifacts will be reconstructed when some adjacent bins of detector don’t work. The ring artifacts severely degrade CT image quality. We present a useful CT ring artifacts reduction based on projection data correction, aiming at estimating the missing data of projection data accurately, thus removing the ring artifacts of CT images. Methods: The method consists of ten steps: 1) Identification of abnormal pixel line in projection sinogram; 2) Linear interpolation within the pixel line of projection sinogram; 3) FBP reconstruction using interpolated projection data; 4) Filtering FBP image using mean filter; 5) Forwarding projection of filtered FBP image; 6) Subtraction forwarded projection from original projection; 7) Linear interpolation of abnormal pixel line area in the subtraction projection; 8) Adding the interpolated subtraction projection on the forwarded projection; 9) FBP reconstruction using corrected projection data; 10) Return to step 4 until the pre-set iteration number is reached. The method is validated on simulated and real data to restore missing projection data and reconstruct ring artifact-free CT images. Results: We have studied impact of amount of dead bins of CT detector on the accuracy of missing data estimation in projection sinogram. For the simulated case with a resolution of 256 by 256 Shepp-Logan phantom, three iterations are sufficient to restore projection data and reconstruct ring artifact-free images when the dead bins rating is under 30%. The dead-bin-induced artifacts are substantially reduced. More iteration number is needed to reconstruct satisfactory images while the rating of dead bins increases. Similar results were found for a real head phantom case. Conclusion: A practical CT image ring artifact correction scheme based on projection data is developed. This method can produce ring artifact-free CT images feasibly and effectively.

  20. SU-F-I-08: CT Image Ring Artifact Reduction Based On Prior Image

    International Nuclear Information System (INIS)

    Yuan, C; Qi, H; Chen, Z; Wu, S; Xu, Y; Zhou, L

    2016-01-01

    Purpose: In computed tomography (CT) system, CT images with ring artifacts will be reconstructed when some adjacent bins of detector don’t work. The ring artifacts severely degrade CT image quality. We present a useful CT ring artifacts reduction based on projection data correction, aiming at estimating the missing data of projection data accurately, thus removing the ring artifacts of CT images. Methods: The method consists of ten steps: 1) Identification of abnormal pixel line in projection sinogram; 2) Linear interpolation within the pixel line of projection sinogram; 3) FBP reconstruction using interpolated projection data; 4) Filtering FBP image using mean filter; 5) Forwarding projection of filtered FBP image; 6) Subtraction forwarded projection from original projection; 7) Linear interpolation of abnormal pixel line area in the subtraction projection; 8) Adding the interpolated subtraction projection on the forwarded projection; 9) FBP reconstruction using corrected projection data; 10) Return to step 4 until the pre-set iteration number is reached. The method is validated on simulated and real data to restore missing projection data and reconstruct ring artifact-free CT images. Results: We have studied impact of amount of dead bins of CT detector on the accuracy of missing data estimation in projection sinogram. For the simulated case with a resolution of 256 by 256 Shepp-Logan phantom, three iterations are sufficient to restore projection data and reconstruct ring artifact-free images when the dead bins rating is under 30%. The dead-bin-induced artifacts are substantially reduced. More iteration number is needed to reconstruct satisfactory images while the rating of dead bins increases. Similar results were found for a real head phantom case. Conclusion: A practical CT image ring artifact correction scheme based on projection data is developed. This method can produce ring artifact-free CT images feasibly and effectively.

  1. TH-CD-207B-10: Effect of CT Reconstruction Filter On Measured Hounsfield Values in Lung Nodules

    Energy Technology Data Exchange (ETDEWEB)

    Little, K; Reiser, I; Sanchez, A; Chung, J; MacMahon, H; Lu, Z [The University of Chicago, Chicago, IL (United States)

    2016-06-15

    Purpose: Measured Hounsfield numbers in CT are used by radiologists to determine the presence of calcium or fat in lung nodules, either of which suggests a benign diagnosis. However, substantial variations in Hounsfield number may arise due to the use of different reconstruction parameters such as the filter/kernel, leading to measurement inaccuracies. This quality improvement project was developed to demonstrate measurement pitfalls and to identify acceptable conditions for incorporating Hounsfield values as a factor in lung nodule diagnosis. Methods: 12 mm-diameter spheres of polyurethane and urethane foam were placed into an anthropomorphic chest phantom, and 10 mm-diameter tubes with varying iodine concentrations were placed into a 16 cm PMMA cylindrical phantom. Additionally, 11 mm-diameter PMMA and HDPE spheres were placed in a 10 cm PMMA cylindrical phantom. Phantoms were scanned at 120 kVp using a Siemens Biograph mCT and on a Philips iCT and reconstructed using various reconstruction filters. Results: For the Siemens system, both sharp kernels and smooth kernels altered the Hounsfield numbers. Hounsfield numbers varied within a range of 8.9 HU for urethane foam and varied within 58.7 HU for polyurethane. The iodine measurements varied up to 37.9 HU for the lowest concentration. For the Philips system, Hounsfield numbers were relatively consistent but were higher for the “Detail” and “Lung Enhanced” filters, varying by 36.9 HU for PMMA and 15.9 HU for HDPE. Conclusion: Reconstruction filters can change the measured Hounsfield numbers of nodular objects, especially with detail-enhancing (sharpening) filters commonly used in lung imaging. Measured values should only be used for diagnostic decision support with filters that have demonstrated accuracy and consistency. While filter accuracy statements are available from manufacturers, radiologists are likely not aware of the extent of potential variations that can occur in a clinical setting.

  2. The diagnostic accuracy of US, CT, MRI and 1H-MRS for the evaluation of hepatic steatosis compared with liver biopsy: a meta-analysis

    International Nuclear Information System (INIS)

    Bohte, Anneloes E.; Werven, Jochem R. van; Bipat, Shandra; Stoker, Jaap

    2011-01-01

    To meta-analyse the diagnostic accuracy of US, CT, MRI and 1 H-MRS for the evaluation of hepatic steatosis. From a comprehensive literature search in MEDLINE, EMBASE, CINAHL and Cochrane (up to November 2009), articles were selected that investigated the diagnostic performance imaging techniques for evaluating hepatic steatosis with histopathology as the reference standard. Cut-off values for the presence of steatosis on liver biopsy were subdivided into four groups: (1) >0, >2 and >5% steatosis; (2) >10, >15 and >20%; (3) >25, >30 and >33%; (4) >50, >60 and >66%. Per group, summary estimates for sensitivity and specificity were calculated. The natural-logarithm of the diagnostic odds ratio (lnDOR) was used as a single indicator of test performance. 46 articles were included. Mean sensitivity estimates for subgroups were 73.3-90.5% (US), 46.1-72.0% (CT), 82.0-97.4% (MRI) and 72.7-88.5% ( 1 H-MRS). Mean specificity ranges were 69.6-85.2% (US), 88.1-94.6% (CT), 76.1-95.3% (MRI) and 92.0-95.7% ( 1 H-MRS). Overall performance (lnDOR) of MRI and 1 H-MRS was better than that for US and CT for all subgroups, with significant differences in groups 1 and 2. MRI and 1 H-MRS can be considered techniques of choice for accurate evaluation of hepatic steatosis. (orig.)

  3. Diagnostic accuracy and utility of coronary CT angiography with consideration of unevaluable results: A systematic review and multivariate Bayesian random-effects meta-analysis with intention to diagnose

    Energy Technology Data Exchange (ETDEWEB)

    Menke, Jan [University Medical Center Goettingen, Institute for Diagnostic and Interventional Radiology, Goettingen (Germany); Kowalski, Joerg [Dr. Lauterbach-Klinik, Department of Cardiology, Bad Liebenstein (Germany)

    2016-02-15

    To meta-analyze diagnostic accuracy, test yield and utility of coronary computed tomography angiography (CCTA) in coronary artery disease (CAD) by an intention-to-diagnose approach with inclusion of unevaluable results. Four databases were searched from 1/2005 to 3/2013 for prospective studies that used 16-320-row or dual-source CTs and provided 3 x 2 patient-level data of CCTA (positive, negative, or unevaluable) versus catheter angiography (positive or negative) for diagnosing ≥50 % coronary stenoses. A Bayesian multivariate 3 x 2 random-effects meta-analysis considered unevaluable CCTAs. Thirty studies (3422 patients) were included. Compared to 16-40 row CT, test yield and accuracy of CCTA has significantly increased with ≥64-row CT (P < 0.05). In ≥64-row CT, about 2.5 % (95 %-CI, 0.9-4.8 %) of diseased patients and 7.5 % (4.5-11.2 %) of non-diseased patients had unevaluable CCTAs. A positive likelihood ratio of 8.9 (6.1-13.5) indicated moderate suitability for identifying CAD. A negative likelihood ratio of 0.022 (0.01-0.04) indicated excellent suitability for excluding CAD. Unevaluable CCTAs had an equivocal likelihood ratio of 0.42 (0.22-0.71). In the utility analysis, CCTA was useful at intermediate pre-test probabilities (16-70 %). CCTA is useful at intermediate CAD pre-test probabilities. Positive CCTAs require verification to confirm CAD, unevaluable CCTAs require alternative diagnostics, and negative CCTAs exclude obstructive CAD with high certainty. (orig.)

  4. Phase- and size-adjusted CT cut-off for differentiating neoplastic lesions from normal colon in contrast-enhanced CT colonography

    International Nuclear Information System (INIS)

    Luboldt, W.; Kroll, M.; Wetter, A.; Vogl, T.J.; Toussaint, T.L.; Hoepffner, N.; Holzer, K.; Kluge, A.

    2004-01-01

    A computed tomography (CT) cut-off for differentiating neoplastic lesions (polyps/carcinoma) from normal colon in contrast-enhanced CT colonography (CTC) relating to the contrast phase and lesion size is determined. CT values of 64 colonic lesions (27 polyps 0 . The slope m was determined by linear regression in the correlation (lesion ∝[xA + (1 - x)V]//H) and the Y-intercept y 0 by the minimal shift of the line needed to maximize the accuracy of separating the colonic wall from the lesions. The CT value of the lesions correlated best with the intermediate phase: 0.4A+ 0.6V(r=0.8 for polyps ≥10 mm, r=0.6 for carcinomas, r=0.4 for polyps <10 mm). The accuracy in the differentiation between lesions and normal colonic wall increased with the height implemented as divisor, reached 91% and was obtained by the dynamic cut-off described by the formula: cut-off(A,V,H) = 1.1[0.4A + 0.6V]/H + 69.8. The CT value of colonic polyps or carcinomas can be increased extrinsically by scanning in the phase in which 0.4A + 0.6V reaches its maximum. Differentiating lesions from normal colon based on CT values is possible in contrast-enhanced CTC and improves when the cut-off is adjusted (normalized) to the contrast phase and lesion size. (orig.)

  5. Role of CT in management of mesothelioma

    International Nuclear Information System (INIS)

    Godwin, J.D.; Rusch, V.W.; Shuman, W.P.

    1987-01-01

    The authors examined the accuracy of CT in determining the extent of disease and its role in the follow-up of patients with malignant pleural mesothelioma. Twenty patients underwent complex CT-anatomic correlation. CT was unable to demonstrate small tumor implants in the chest wall (four cases), upper abdomen (two cases), or contralateral hemidiaphragm (one case). Occasionally there was difficulty distinguishing tumor invasion of soft tissues from simple contiguity, benign from malignant nodes, and recurrent tumor from surgical changes. In six of eight treated patients, CT demonstrated recurrence -8 months before signs or symptoms appeared. Despite its limitations, CT is essential in the initial evaluation and follow-up of patients with mesothelioma

  6. FDG-PET/CT and FLT-PET/CT for differentiating between lipid-poor benign and malignant adrenal tumours

    Energy Technology Data Exchange (ETDEWEB)

    Nakajo, Masatoyo; Jinguji, Megumi; Fukukura, Yoshihiko; Nakabeppu, Yoshiaki; Nakamura, Fumihiko; Yoshiura, Takashi [Kagoshima University, Department of Radiology, Graduate School of Medical and Dental Sciences, Kagoshima (Japan); Kajiya, Yoriko; Tani, Atushi; Nakajo, Masayuki [Nanpuh Hospital, Department of Radiology, Kagoshima (Japan); Arimura, Hiroshi; Nishio, Yoshihiko [Kagoshima University, Department of Diabetes and Endocrine Medicine, Graduate School of Medical and Dental Sciences, Kagoshima (Japan)

    2015-12-15

    To compare F-18-fluorodeoxyglucose (FDG) and F-18-fluorothymidine (FLT) PET/CT examinations for differentiating between benign and malignant adrenal tumours. Thirty lipid-poor benign and 11 malignant tumours of 40 patients were included. FDG- and FLT-based indices including visual score, maximum standardized uptake value (SUVmax) and FDG adrenal lesion/liver SUVmax (A/L SUVmax) or FLT adrenal lesion/back muscle SUVmax (A/B SUVmax) ratio were compared between benign and malignant tumours using the Mann-Whitney's U or Wilcoxon signed-rank test, and their diagnostic performances were evaluated by means of the area under the curve (AUC) values derived from the receiver operating characteristic analysis. All indices were significantly higher in malignant than benign tumours on both images (p < 0.05 each). On FDG-PET/CT, the sensitivity, specificity, and accuracy were 91 %, 63 % and 71 % for visual score, 91 %, 67 % and 73 % for SUVmax, and 100 %, 70 % and 78 % for A/L SUVmax ratio, respectively. On FLT-PET/CT, they were 100 %, 97 % and 98 % for visual score, SUVmax and A/B SUVmax ratio, respectively. All FLT indices were significantly higher than those of FDG in AUC (p < 0.05 each). FLT-PET/CT may be superior to FDG-PET/CT in differentiating lipid-poor benign from malignant adrenal tumours because of higher specificity and accuracy. (orig.)

  7. Tracker-on-C for cone-beam CT-guided surgery: evaluation of geometric accuracy and clinical applications

    Science.gov (United States)

    Reaungamornrat, S.; Otake, Y.; Uneri, A.; Schafer, S.; Mirota, D. J.; Nithiananthan, S.; Stayman, J. W.; Khanna, A. J.; Reh, D. D.; Gallia, G. L.; Taylor, R. H.; Siewerdsen, J. H.

    2012-02-01

    Conventional surgical tracking configurations carry a variety of limitations in line-of-sight, geometric accuracy, and mismatch with the surgeon's perspective (for video augmentation). With increasing utilization of mobile C-arms, particularly those allowing cone-beam CT (CBCT), there is opportunity to better integrate surgical trackers at bedside to address such limitations. This paper describes a tracker configuration in which the tracker is mounted directly on the Carm. To maintain registration within a dynamic coordinate system, a reference marker visible across the full C-arm rotation is implemented, and the "Tracker-on-C" configuration is shown to provide improved target registration error (TRE) over a conventional in-room setup - (0.9+/-0.4) mm vs (1.9+/-0.7) mm, respectively. The system also can generate digitally reconstructed radiographs (DRRs) from the perspective of a tracked tool ("x-ray flashlight"), the tracker, or the C-arm ("virtual fluoroscopy"), with geometric accuracy in virtual fluoroscopy of (0.4+/-0.2) mm. Using a video-based tracker, planning data and DRRs can be superimposed on the video scene from a natural perspective over the surgical field, with geometric accuracy (0.8+/-0.3) pixels for planning data overlay and (0.6+/-0.4) pixels for DRR overlay across all C-arm angles. The field-of-view of fluoroscopy or CBCT can also be overlaid on real-time video ("Virtual Field Light") to assist C-arm positioning. The fixed transformation between the x-ray image and tracker facilitated quick, accurate intraoperative registration. The workflow and precision associated with a variety of realistic surgical tasks were significantly improved using the Tracker-on-C - for example, nearly a factor of 2 reduction in time required for C-arm positioning, reduction or elimination of dose in "hunting" for a specific fluoroscopic view, and confident placement of the x-ray FOV on the surgical target. The proposed configuration streamlines the integration of C

  8. How to perform dosimetry with Optical CT

    International Nuclear Information System (INIS)

    Wuu, Cheng-Shie; Xu, Y

    2010-01-01

    Both polymer gels and PRESAGE radiochromic solid dosimeter, in conjunction with optical CT scanning system, have been employed to measure 3-D dose distribution. The 3-D dose maps obtained from these systems can provide a useful tool for dose verification on complex treatments such as IMRT, radiosurgery, and RapidArc. These complex treatments present high dose gradient regions in the boundaries between the target and the surrounding critical organs. Dose accuracy in these areas can be critical, and may affect the treatment. There is a pressing need for a dosimeter that allows for accurate determination of 3-D dose distribution with high spatial resolution. In this paper, performance of polymer gels and PRESAGE dosimeter with optical CT scanning is reviewed and evaluated in terms of their sensitivity calibration, irradiation, optimization of scanning procedures, precision, and accuracy. Clinical applications of optical-CT dosimetry are presented.

  9. Accuracy and Radiation Dose Reduction of Limited-Range CT in the Evaluation of Acute Appendicitis in Pediatric Patients.

    Science.gov (United States)

    Jin, Michael; Sanchez, Thomas R; Lamba, Ramit; Fananapazir, Ghaneh; Corwin, Michael T

    2017-09-01

    The purpose of this article is to determine the accuracy and radiation dose reduction of limited-range CT prescribed from the top of L2 to the top of the pubic symphysis in children with suspected acute appendicitis. We performed a retrospective study of 210 consecutive pediatric patients from December 11, 2012, through December 11, 2014, who underwent abdominopelvic CT for suspected acute appendicitis. Two radiologists independently reviewed the theoretic limited scans from the superior L2 vertebral body to the top of the pubic symphysis, to assess for visualization of the appendix, acute appendicitis, alternative diagnoses, and incidental findings. Separately, the same parameters were assessed on the full scan by the same two reviewers. Whole-body effective doses were determined for the full- and limited-range scans and were compared using the paired t test. The appendix or entire cecum was visualized on the limited scan in all cases, and no cases of acute appendicitis were missed on the simulated limited scan compared with the full scan. Two alternative diagnoses were missed with the limited scan: one case of hydronephrosis and one of acute acalculous cholecystitis. The mean effective dose for the original scan was 5.6 mSv and that for the simulated limited scan was 3.0 mSv, resulting in a dose reduction of 46.4% (p appendicitis and reduces the dose by approximately 46%.

  10. Accuracy of stereolithographic models of human anatomy

    International Nuclear Information System (INIS)

    Barker, T.M.; Earwaker, W.J.S.; Lisle, D.A.

    1994-01-01

    A study was undertaken to determine the dimensional accuracy of anatomical replicas derived from X-ray 3D computed tomography (CT) images and produced using the rapid prototyping technique of stereolithography (SLA). A dry bone skull and geometric phantom were scanned, and replicas were produced. Distance measurements were obtained to compare the original objects and the resulting replicas. Repeated measurements between anatomical landmarks were used for comparison of the original skull and replica. Results for the geometric phantom demonstrate a mean difference of +0.47mm, representing an accuracy of 97.7-99.12%. Measurements of the skull produced a range of absolute differences (maximum +4.62mm, minimum +0.1mm, mean +0.85mm). These results support the use of SLA models of human anatomical structures in such areas as pre-operative planning of complex surgical procedures. For applications where higher accuracy is required, improvements can be expected by utilizing smaller pixel resolution in the CT images. Stereolithographic models can now be confidently employed as accurate, three-dimensional replicas of complex, anatomical structures. 14 refs., 2 tabs., 8 figs

  11. Measurement methods and accuracy in copy number variation: failure to replicate associations of beta-defensin copy number with Crohn's disease

    Science.gov (United States)

    Aldhous, Marian C.; Abu Bakar, Suhaili; Prescott, Natalie J.; Palla, Raquel; Soo, Kimberley; Mansfield, John C.; Mathew, Christopher G.; Satsangi, Jack; Armour, John A.L.

    2010-01-01

    The copy number variation in beta-defensin genes on human chromosome 8 has been proposed to underlie susceptibility to inflammatory disorders, but presents considerable challenges for accurate typing on the scale required for adequately powered case–control studies. In this work, we have used accurate methods of copy number typing based on the paralogue ratio test (PRT) to assess beta-defensin copy number in more than 1500 UK DNA samples including more than 1000 cases of Crohn's disease. A subset of 625 samples was typed using both PRT-based methods and standard real-time PCR methods, from which direct comparisons highlight potentially serious shortcomings of a real-time PCR assay for typing this variant. Comparing our PRT-based results with two previous studies based only on real-time PCR, we find no evidence to support the reported association of Crohn's disease with either low or high beta-defensin copy number; furthermore, it is noteworthy that there are disagreements between different studies on the observed frequency distribution of copy number states among European controls. We suggest safeguards to be adopted in assessing and reporting the accuracy of copy number measurement, with particular emphasis on integer clustering of results, to avoid reporting of spurious associations in future case–control studies. PMID:20858604

  12. Measurement methods and accuracy in copy number variation: failure to replicate associations of beta-defensin copy number with Crohn's disease.

    Science.gov (United States)

    Aldhous, Marian C; Abu Bakar, Suhaili; Prescott, Natalie J; Palla, Raquel; Soo, Kimberley; Mansfield, John C; Mathew, Christopher G; Satsangi, Jack; Armour, John A L

    2010-12-15

    The copy number variation in beta-defensin genes on human chromosome 8 has been proposed to underlie susceptibility to inflammatory disorders, but presents considerable challenges for accurate typing on the scale required for adequately powered case-control studies. In this work, we have used accurate methods of copy number typing based on the paralogue ratio test (PRT) to assess beta-defensin copy number in more than 1500 UK DNA samples including more than 1000 cases of Crohn's disease. A subset of 625 samples was typed using both PRT-based methods and standard real-time PCR methods, from which direct comparisons highlight potentially serious shortcomings of a real-time PCR assay for typing this variant. Comparing our PRT-based results with two previous studies based only on real-time PCR, we find no evidence to support the reported association of Crohn's disease with either low or high beta-defensin copy number; furthermore, it is noteworthy that there are disagreements between different studies on the observed frequency distribution of copy number states among European controls. We suggest safeguards to be adopted in assessing and reporting the accuracy of copy number measurement, with particular emphasis on integer clustering of results, to avoid reporting of spurious associations in future case-control studies.

  13. CT diagnosis of cystic ovarian lesions

    International Nuclear Information System (INIS)

    Iio, Kazuto; Shinmura, Ryoji; Arima, Naomi; Yamada, Eiichiro; Ohkubo, Koichi; Nagata, Yukihiro

    1985-01-01

    CT was undertaken and CT numbers were measured in 47 patients with cystic ovarian lesions. CT features particularly for chocolate cyst revealed the uniform thickness of the whole cystic wall, findings suggesting adhesion to the surrounding organs, circular or oval shape, and higher CT numbers within the cyst than those in the other cystic ovarian lesions. However, because these features are not always observed in cases of chocolate cyst, one should not rely solely on CT findings in the diagnosis of chocolate cyst. (Namekawa, K.)

  14. Xanthogranulomatous cholecystitis: Diagnostic performance of CT to differentiate from gallbladder cancer

    Energy Technology Data Exchange (ETDEWEB)

    Goshima, Satoshi, E-mail: gossy@par.odn.ne.j [Department of Radiology, University of Pittsburgh Medical Center, Lothrop St., Pittsburgh, PA 15213 (United States); Department of Radiology, Gifu University School of Medicine, 1-1 Yanagido, Gifu 501-1194 (Japan); Chang, Samuel; Wang, Jin Hong [Department of Radiology, University of Pittsburgh School of Medicine, 3362 Fifth Ave., Pittsburgh, PA15213 (United States); Kanematsu, Masayuki [Department of Radiology, Gifu University School of Medicine, 1-1 Yanagido, Gifu 501-1194 (Japan); Department of Radiology Services, Gifu University School of Medicine, 1-1- Yanagido, Gifu 501-1194 (Japan); Bae, Kyongtae T. [Department of Radiology, University of Pittsburgh School of Medicine, 3362 Fifth Ave., Pittsburgh, PA15213 (United States); Federle, Michael P. [Department of Radiology, University of Pittsburgh Medical Center, Lothrop St., Pittsburgh, PA 15213 (United States); Department of Radiology, Stanford University Medical Center, 300 Pasteur Drive, Stanford, CA 94305-5105 (United States)

    2010-06-15

    Purpose: To retrospectively evaluate CT findings of xanthogranulomatous cholecystitis (XGC) and to measure diagnostic performance for distinguishing it from gallbladder (GB) cancer. Methods and materials: Our institutional review board approved this retrospective study. Three blinded radiologists, first independently and then in consensus, retrospectively evaluated postcontrast CT images of 35 patients with histopathologically proved XGC and GB cancer, all of whom subsequently had cholecystectomy. These included 18 patients with XGC (13 male, 5 female; age range, 35-84, mean 63 years) and 17 with GB cancer (6 male, 11 female; age range, 45-95, mean 69). Differences in CT findings between XGC and GB cancer and diagnostic performances for each CT finding were calculated. Sensitivity, specificity, and accuracy were calculated for each radiologist and observer performance was also determined by receiver-operating-characteristic curve analysis. Results: Five CT findings showed significant differences between XGC and GB cancer. Sensitivity, specificity, and accuracy of each finding for the differentiation of XGC were 89%, 65%, 77% with diffuse GB wall thickening, 67%, 82%, 74% with a continuous mucosal line, 61%, 71%, 66% with intra-mural hypo-attenuated nodules, 72%, 77%, 74% with absence of macroscopic hepatic invasion, and 67%, 71%, 69% with absence of intra-hepatic bile duct dilatation, respectively. When at least three of these five CT findings were observed in combination, sensitivity, specificity, and accuracy were 83%, 100% and 91%, respectively. Sensitivities, specificities and Az values for the differentiation of XGC from GB cancer were 83%, 88%, 0.94 for reader 1, 78%, 88%, 0.93 for reader 2, and 78%, 82%, 0.84 for reader 3. Conclusions: The combination of three of the five CT findings that are common with XGC can provide excellent accuracy for the differentiation of XGC and GB cancer.

  15. Dual-energy CT perfusion and angiography in chronic thromboembolic pulmonary hypertension: diagnostic accuracy and concordance with radionuclide scintigraphy

    Energy Technology Data Exchange (ETDEWEB)

    Dournes, Gael; Verdier, Damien; Montaudon, Michel; Laurent, Francois; Lederlin, Mathieu [Hopital Haut-Leveque, CHU Bordeaux, Department of Medical Imaging, Pessac (France); University Bordeaux Segalen, Bordeaux Cedex (France); Bullier, Eric; Riviere, Annalisa [Hopital Haut-Leveque, CHU Bordeaux, Department of Nuclear Medicine, Pessac (France); Dromer, Claire [Hopital Haut-Leveque, CHU Bordeaux, Department of Respiratory Diseases, Pessac (France); Picard, Francois [Hopital Haut-Leveque, CHU Bordeaux, Department of Cardiology, Pessac (France); Billes, Marc-Alain [Hopital Haut-Leveque, CHU Bordeaux, Department of Cardiac Surgery, Pessac (France); Corneloup, Olivier [Hopital Haut-Leveque, CHU Bordeaux, Department of Medical Imaging, Pessac (France)

    2014-01-15

    To evaluate the diagnostic accuracy of dual-energy computed tomography (DECT) perfusion and angiography versus ventilation/perfusion (V/Q) scintigraphy in chronic thromboembolic pulmonary hypertension (CTEPH), and to assess the per-segment concordance rate of DECT and scintigraphy. Forty consecutive patients with proven pulmonary hypertension underwent V/Q scintigraphy and DECT perfusion and angiography. Each imaging technique was assessed for the location of segmental defects. Diagnosis of CTEPH was established when at least one segmental perfusion defect was detected by scintigraphy. Diagnostic accuracy of DECT perfusion and angiography was assessed and compared with scintigraphy. In CTEPH patients, the per-segment concordance between scintigraphy and DECT perfusion/angiography was calculated. Fourteen patients were diagnosed with CTEPH and 26 with other aetiologies. DECT perfusion and angiography correctly identified all CTEPH patients with sensitivity/specificity values of 1/0.92 and 1/0.93, respectively. At a segmental level, DECT perfusion showed moderate agreement (κ = 0.44) with scintigraphy. Agreement between CT angiography and scintigraphy ranged from fair (κ = 0.31) to slight (κ = 0.09) depending on whether completely or partially occlusive patterns were considered, respectively. Both DECT perfusion and angiography show satisfactory performance for the diagnosis of CTEPH. DECT perfusion is more accurate than angiography at identifying the segmental location of abnormalities. (orig.)

  16. Dual-energy CT perfusion and angiography in chronic thromboembolic pulmonary hypertension: diagnostic accuracy and concordance with radionuclide scintigraphy

    International Nuclear Information System (INIS)

    Dournes, Gael; Verdier, Damien; Montaudon, Michel; Laurent, Francois; Lederlin, Mathieu; Bullier, Eric; Riviere, Annalisa; Dromer, Claire; Picard, Francois; Billes, Marc-Alain; Corneloup, Olivier

    2014-01-01

    To evaluate the diagnostic accuracy of dual-energy computed tomography (DECT) perfusion and angiography versus ventilation/perfusion (V/Q) scintigraphy in chronic thromboembolic pulmonary hypertension (CTEPH), and to assess the per-segment concordance rate of DECT and scintigraphy. Forty consecutive patients with proven pulmonary hypertension underwent V/Q scintigraphy and DECT perfusion and angiography. Each imaging technique was assessed for the location of segmental defects. Diagnosis of CTEPH was established when at least one segmental perfusion defect was detected by scintigraphy. Diagnostic accuracy of DECT perfusion and angiography was assessed and compared with scintigraphy. In CTEPH patients, the per-segment concordance between scintigraphy and DECT perfusion/angiography was calculated. Fourteen patients were diagnosed with CTEPH and 26 with other aetiologies. DECT perfusion and angiography correctly identified all CTEPH patients with sensitivity/specificity values of 1/0.92 and 1/0.93, respectively. At a segmental level, DECT perfusion showed moderate agreement (κ = 0.44) with scintigraphy. Agreement between CT angiography and scintigraphy ranged from fair (κ = 0.31) to slight (κ = 0.09) depending on whether completely or partially occlusive patterns were considered, respectively. Both DECT perfusion and angiography show satisfactory performance for the diagnosis of CTEPH. DECT perfusion is more accurate than angiography at identifying the segmental location of abnormalities. (orig.)

  17. The accuracy of CT-based inhomogeneity corrections and in vivo dosimetry for the treatment of lung cancer

    International Nuclear Information System (INIS)

    Essers, M.; Lanson, J.H.; Leunens, G.; Schnabel, T.; Mijnheer, B.J.

    1995-01-01

    Purpose: To determine the accuracy of dose calculations based on CT-densities for lung cancer patients irradiated with an anterioposterior parallel-opposed treatment technique and to evaluate, for this technique, the use of diodes and an Electronic Portal Imaging Device (EPID) for absolute exit dose and relative transmission dose verification, respectively. Materials and methods: Dose calculations were performed using a 3-dimensional treatment planning system, using CT-densities or assuming the patient to be water-equivalent. A simple inhomogeneity correction model was used to take CT-densities into account. For 22 patients, entrance and exit dose calculations at the central beam axis and at several off-axis positions were compared with diode measurements. For 12 patients, diode exit dose measurements and exit dose calculations were compared with EPID transmission dose values. Results: Using water-equivalent calculations, the actual exit dose value under lung was, on average, underestimated by 30%, with an overall spread of 10% (1 SD) in the ratio of measurement and calculation. Using inhomogeneity corrections, the exit dose was, on average, overestimated by 4%, with an overall spread of 6% (1 SD). Only 2% of the average deviation was due to the inhomogeneity correction model. The other 2% resulted from a small inaccuracy in beam fit parameters and the fact that lack of backscatter is not taken into account by the calculation model. Organ motion, resulting from the ventilatory or cardiac cycle, caused an estimated uncertainty in calculated exit dose of 2.5% (1 SD). The most important reason for the large overall spread was, however, the inaccuracy involved in point measurements, of about 4% (1 SD), which resulted from the systematic and random deviation in patient set-up and therefore in the diode position with respect to patient anatomy. Transmission and exit dose values agreed with an average difference of 1.1%. Transmission dose profiles also showed good

  18. Value of PET-CT and PET-CT combined with lung VCAR software in the diagnosis of hilar area lymph nodes of non-small cell lung cancer

    International Nuclear Information System (INIS)

    Yu Lijuan; Li Yingci; Wang Wenzhi; Wang Xin; Lu Pei'ou; Tian Mohan

    2012-01-01

    Objective: To explore the diagnostic value of PET-CT and PET-CT combined with lung volume computed assisted reading (Lung VCAR) software in hilar area lymph nodes. Methods: Preoperative whole body PET-CT imaging was performed in 49 patients who were highly suspicious of non-small cell lung cancer. PET-CT images of the hilar area lymph nodes and the PET-CT images of the hilar area lymph nodes from Lung VCAR software were evaluated by two experienced doctors, and then compared with the pathological diagnosis. Results: There was no significant difference between the CT values of benign and malignant lymph nodes (t=-1.40, P>0.05). But a significant difference was existed between the benign and malignant hilar lymph nodes with the density visual analysis, the lymph short diameter and the maximum of standardized uptake value (SUV max ) (χ 2 =30.37, 27.40, 20.06, all P<0.05). The sensibility,specificity and accuracy of PET-CT in diagnosis of the hilar area lymph nodes were 76.5%, 90.7%, 88.3% respectively, and the accuracy of the diagnosis was significantly higher than that of CT and PET alone (χ 2 =15.27, P<0.05) using the lymph short diameter ≥1 cm of CT, the density of lymph node is equal to (slightly lower than) the same layer vascular density and the lymph node SUV max ≥2.5 of PET as the diagnostic criteria. One hundred and three hilar area lymph nodes were diagnosed by PET-CT and four nodes were not hilar lymph nodes proved by the Lung VCAR software (3 hilar vascular uptake,1 bronchial cartilage). Conclusion: The methods of PET-CT lymph visual density analysis plus lymph node diameter and SUV max had a high diagnostic accuracy of non-small cell lung hilar lymph. For the PET-CT,the pulmonary vascular uptake was the main cause affecting the discrimination of hilar lymph nodes,while Lung VCAR software was helpful to diagnosis. (authors)

  19. Influence of heart rate on diagnostic accuracy and image quality of 16-slice CT coronary angiography: comparison of multisegment and halfscan reconstruction approaches

    Energy Technology Data Exchange (ETDEWEB)

    Dewey, Marc; Teige, Florian; Hamm, Bernd [Charite - Universitaetsmedizin Berlin, Humboldt-Universitaet zu Berlin, Department of Radiology, Chariteplatz 1, P.O. Box 10098, Berlin (Germany); Laule, Michael [Department of Cardiology, Charite, Berlin (Germany)

    2007-11-15

    The lower the heart rate the better image quality in multislice computed tomography (MSCT) coronary angiography. We prospectively assessed the influence of heart rate on per-patient diagnostic accuracy and image quality of MSCT coronary angiography and compared adaptive multisegment and standard halfscan reconstruction. A consecutive cohort of 126 patients scheduled to undergo conventional coronary angiography was examined with 16-slice CT. For all heart rate groups, per-patient diagnostic accuracy was significantly higher for multisegment than halfscan reconstruction with values of 95 vs. 79% (p < 0.05, <65 bpm, 38 patients), 85 vs. 66% (p < 0.05, 65-74 bpm, 47 patients), and 78% vs. 41% (p < 0.001, >74 bpm, 41 patients). Differences in diagnostic accuracy between adjacent heart rate groups were only significant for halfscan reconstruction for the comparison between the 65-74 and >74 bpm group (p < 0.05). The vessel lengths free of motion artifacts were significantly longer with multisegment reconstruction in all heart rate groups and for all coronary arteries (p < 0.005). For noninvasive MSCT coronary angiography, both per-patient diagnostic accuracy and image quality decline with increasing heart rate, and multisegment reconstruction at high heart rates yields similar results as standard halfscan reconstruction at low heart rates. (orig.)

  20. Standardized CT examination of the multitraumatized patient

    Energy Technology Data Exchange (ETDEWEB)

    Leidner, B. [Department of Diagnostic Radiology, Oskarshamn Hospital (Sweden)]|[Department of Diagnostic Radiology, Huddinge University Hospital (Sweden); Adiels, M. [Department of Diagnostic Radiology, Oskarshamn Hospital (Sweden); Aspelin, P. [Department of Diagnostic Radiology, Huddinge University Hospital (Sweden); Gullstrand, P.; Wallen, S. [Department of Surgery, Oskarshamn Hospital (Sweden)

    1998-12-01

    The aim of this study was to develop a standardized non-helical-CT protocol including head, body and proximal extremities in order to achieve a good time efficiency and diagnostic accuracy in the initial radiological evaluation of the multitraumatized patient. A total of 111 circulatory stable blunt trauma patients, brought in to a trauma level II-III hospital, were examined according to a standardized CT protocol. After examining the head with contiguous 10-mm slices without IV contrast medium injection, the trunk was examined with 10-mm slices every 30 mm through thorax-abdomen-pelvis with IV contrast medium enhancement (occasionally modified). All data in the medical reports were collected and used as ``end-point``, and the outcome of the CT examination was compared with this final diagnosis. Mean examination time was 20 min (range 12-32 min). In total, 55 head injuries, 89 thoracic injuries, 27 abdominal/pelvic injuries and 62 fractures were found. Computed tomography correctly identified the injuries, except one brain stem injury, one contusion/rupture of the heart, one hepatic injury, two intestinal injuries, eight vertebral injuries and one joint dislocation. A standardized non-helical-CT examination of the head and body may be achieved in 20 min. Its diagnostic accuracy was high, except for vertebral column injuries, which is why we recommend it as the method of choice for initial radiological examination of multitraumatized patients. When available, helical scanning would improve both examination speed and accuracy. (orig.) With 6 figs., 40 refs.

  1. Implementing an ultrasound-based protocol for diagnosingappendicitis while maintaining diagnostic accuracy

    International Nuclear Information System (INIS)

    Van Atta, Angela J.; Baskin, Henry J.; Maves, Connie K.; Dansie, David M.; Rollins, Michael D.; Bolte, Robert G.; Mundorff, Michael B.; Andrews, Seth P.

    2015-01-01

    The use of ultrasound to diagnose appendicitis in children is well-documented but not universally employed outside of pediatric academic centers, especially in the United States. Various obstacles make it difficult for institutions and radiologists to abandon a successful and accurate CT-based imaging protocol in favor of a US-based protocol. To describe how we overcame barriers to implementing a US-based appendicitis protocol among a large group of nonacademic private-practice pediatric radiologists while maintaining diagnostic accuracy and decreasing medical costs. A multidisciplinary team of physicians (pediatric surgery, pediatric emergency medicine and pediatric radiology) approved an imaging protocol using US as the primary modality to evaluate suspected appendicitis with CT for equivocal cases. The protocol addressed potential bias against US and accommodated for institutional limitations of radiologist and sonographer experience and availability. Radiologists coded US reports according to the probability of appendicitis. Radiology reports were compared with clinical outcomes to assess diagnostic accuracy. During the study period, physicians from each group were apprised of the interim US protocol accuracy results. Problematic cases were discussed openly. A total of 512 children were enrolled and underwent US for evaluation of appendicitis over a 30-month period. Diagnostic accuracy was comparable to published results for combined US/CT protocols. Comparing the first 12 months to the last 12 months of the study period, the proportion of children achieving an unequivocal US result increased from 30% (51/169) to 53% (149/282) and the proportion of children undergoing surgery based solely on US findings increased from 55% (23/42) to 84% (92/109). Overall, 63% (325/512) of patients in the protocol did not require a CT. Total patient costs were reduced by $30,182 annually. We overcame several barriers to implementing a US protocol. During the study period our

  2. Implementing an ultrasound-based protocol for diagnosingappendicitis while maintaining diagnostic accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Van Atta, Angela J. [University of Utah School of Medicine, Salt Lake City, UT (United States); Baskin, Henry J.; Maves, Connie K.; Dansie, David M. [Primary Children' s Hospital, Department of Radiology, Salt Lake City, UT (United States); Rollins, Michael D. [University of Utah School of Medicine, Department of Surgery, Division of Pediatric Surgery, Salt Lake City, UT (United States); Bolte, Robert G. [University of Utah School of Medicine, Department of Pediatrics, Division of Pediatric Emergency Medicine, Salt Lake City, UT (United States); Mundorff, Michael B.; Andrews, Seth P. [Primary Children' s Hospital, Systems Improvement, Salt Lake City, UT (United States)

    2015-05-01

    The use of ultrasound to diagnose appendicitis in children is well-documented but not universally employed outside of pediatric academic centers, especially in the United States. Various obstacles make it difficult for institutions and radiologists to abandon a successful and accurate CT-based imaging protocol in favor of a US-based protocol. To describe how we overcame barriers to implementing a US-based appendicitis protocol among a large group of nonacademic private-practice pediatric radiologists while maintaining diagnostic accuracy and decreasing medical costs. A multidisciplinary team of physicians (pediatric surgery, pediatric emergency medicine and pediatric radiology) approved an imaging protocol using US as the primary modality to evaluate suspected appendicitis with CT for equivocal cases. The protocol addressed potential bias against US and accommodated for institutional limitations of radiologist and sonographer experience and availability. Radiologists coded US reports according to the probability of appendicitis. Radiology reports were compared with clinical outcomes to assess diagnostic accuracy. During the study period, physicians from each group were apprised of the interim US protocol accuracy results. Problematic cases were discussed openly. A total of 512 children were enrolled and underwent US for evaluation of appendicitis over a 30-month period. Diagnostic accuracy was comparable to published results for combined US/CT protocols. Comparing the first 12 months to the last 12 months of the study period, the proportion of children achieving an unequivocal US result increased from 30% (51/169) to 53% (149/282) and the proportion of children undergoing surgery based solely on US findings increased from 55% (23/42) to 84% (92/109). Overall, 63% (325/512) of patients in the protocol did not require a CT. Total patient costs were reduced by $30,182 annually. We overcame several barriers to implementing a US protocol. During the study period our

  3. Semi-automated scoring of pulmonary emphysema from X-ray CT: Trainee reproducibility and accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Owrangi, Amir M., E-mail: aowrangi@robarts.ca [Imaging Research Laboratories, Robarts Research Institute, London (Canada); Entwistle, Brandon, E-mail: Brandon.Entwistle@londonhospitals.ca; Lu, Andrew, E-mail: Andrew.Lu@londonhospitals.ca; Chiu, Jack, E-mail: Jack.Chiu@londonhospitals.ca; Hussain, Nabil, E-mail: Nabil.Hussain@londonhospitals.ca; Etemad-Rezai, Roya, E-mail: Roya.EtemadRezai@lhsc.on.ca; Parraga, Grace, E-mail: gparraga@robarts.ca [Imaging Research Laboratories, Robarts Research Institute, London (Canada); Graduate Program in Biomedical Engineering, Department of Medical Imaging, Department of Medical Biophysics, The University of Western Ontario, London (Canada)

    2013-11-01

    Objective: We developed a semi-automated tool to quantify emphysema from thoracic X-ray multi-detector (64-slice) computed tomography (CT) for training purposes and multi-reader studies. Materials and Methods: Thoracic X-ray CT was acquired in 93 ex-smokers, who were evaluated by six trainees with little or no expertise (trainees) and a single experienced thoracic radiologist (expert). A graphic user interface (GUI) was developed for emphysema quantification based on the percentile of lung where a score of 0 = no abnormalities, 1 = 1–25%, 2 = 26–50%, 3 = 51–75% and 4 = 76–100% for each lung side/slice. Trainees blinded to subject characteristics scored randomized images twice; accuracy was determined by comparison to expert scores, density histogram 15th percentile (HU{sub 15}), relative area at −950 HU (RA{sub 950}), low attenuation clusters at −950 HU (LAC{sub 950}), −856 HU (LAC{sub 856}) and the diffusing capacity for carbon monoxide (DL{sub CO%pred}). Intra- and inter-observer reproducibility was evaluated using coefficients-of-variation (COV), intra-class (ICC) and Pearson correlations. Results: Trainee–expert correlations were significant (r = 0.85–0.97, p < 0.0001) and a significant trainee bias (0.15 ± 0.22) was observed. Emphysema score was correlated with RA{sub 950} (r = 0.88, p < 0.0001), HU{sub 15} (r = −0.77, p < 0.0001), LAC{sub 950} (r = 0.76, p < 0.0001), LAC{sub 856} (r = 0.74, p = 0.0001) and DL{sub CO%pred} (r = −0.71, p < 0.0001). Intra-observer reproducibility (COV = 4–27%; ICC = 0.75–0.94) was moderate to high for trainees; intra- and inter-observer COV were negatively and non-linearly correlated with emphysema score. Conclusion: We developed a GUI for rapid and interactive emphysema scoring that allows for comparison of multiple readers with clinical and radiological standards.

  4. Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT.

    Science.gov (United States)

    Parsa, Azin; Ibrahim, Norliza; Hassan, Bassam; van der Stelt, Paul; Wismeijer, Daniel

    2015-01-01

    The first purpose of this study was to analyze the correlation between bone volume fraction (BV/TV) and calibrated radiographic bone density Hounsfield units (HU) in human jaws, derived from micro-CT and multislice computed tomography (MSCT), respectively. The second aim was to assess the accuracy of cone beam computed tomography (CBCT) in evaluating trabecular bone density and microstructure using MSCT and micro-CT, respectively, as reference gold standards. Twenty partially edentulous human mandibular cadavers were scanned by three types of CT modalities: MSCT (Philips, Best, the Netherlands), CBCT (3D Accuitomo 170, J Morita, Kyoto, Japan), and micro-CT (SkyScan 1173, Kontich, Belgium). Image analysis was performed using Amira (v4.1, Visage Imaging Inc., Carlsbad, CA, USA), 3Diagnosis (v5.3.1, 3diemme, Cantu, Italy), Geomagic (studio(®) 2012, Morrisville, NC, USA), and CTAn (v1.11, SkyScan). MSCT, CBCT, and micro-CT scans of each mandible were matched to select the exact region of interest (ROI). MSCT HU, micro-CT BV/TV, and CBCT gray value and bone volume fraction of each ROI were derived. Statistical analysis was performed to assess the correlations between corresponding measurement parameters. Strong correlations were observed between CBCT and MSCT density (r = 0.89) and between CBCT and micro-CT BV/TV measurements (r = 0.82). Excellent correlation was observed between MSCT HU and micro-CT BV/TV (r = 0.91). However, significant differences were found between all comparisons pairs (P micro-CT BV/TV (P = 0.147). An excellent correlation exists between bone volume fraction and bone density as assessed on micro-CT and MSCT, respectively. This suggests that bone density measurements could be used to estimate bone microstructural parameters. A strong correlation also was found between CBCT gray values and BV/TV and their gold standards, suggesting the potential of this modality in bone quality assessment at implant site. © 2013 John Wiley & Sons A/S. Published by

  5. SU-C-207A-03: Development of Proton CT Imaging System Using Thick Scintillator and CCD Camera

    Energy Technology Data Exchange (ETDEWEB)

    Tanaka, S; Uesaka, M [The University of Tokyo, Tokyo (Japan); Nishio, T; Tsuneda, M [Hiroshima University, Hiroshima (Japan); Matsushita, K [Rikkyo University, Tokyo (Japan); Kabuki, S [Tokai University, Isehara (Japan)

    2016-06-15

    Purpose: In the treatment planning of proton therapy, Water Equivalent Length (WEL), which is the parameter for the calculation of dose and the range of proton, is derived by X-ray CT (xCT) image and xCT-WEL conversion. However, about a few percent error in the accuracy of proton range calculation through this conversion has been reported. The purpose of this study is to construct a proton CT (pCT) imaging system for an evaluation of the error. Methods: The pCT imaging system was constructed with a thick scintillator and a cooled CCD camera, which acquires the two-dimensional image of integrated value of the scintillation light toward the beam direction. The pCT image is reconstructed by FBP method using a correction between the light intensity and residual range of proton beam. An experiment for the demonstration of this system was performed with 70-MeV proton beam provided by NIRS cyclotron. The pCT image of several objects reconstructed from the experimental data was evaluated quantitatively. Results: Three-dimensional pCT images of several objects were reconstructed experimentally. A finestructure of approximately 1 mm was clearly observed. The position resolution of pCT image was almost the same as that of xCT image. And the error of proton CT pixel value was up to 4%. The deterioration of image quality was caused mainly by the effect of multiple Coulomb scattering. Conclusion: We designed and constructed the pCT imaging system using a thick scintillator and a CCD camera. And the system was evaluated with the experiment by use of 70-MeV proton beam. Three-dimensional pCT images of several objects were acquired by the system. This work was supported by JST SENTAN Grant Number 13A1101 and JSPS KAKENHI Grant Number 15H04912.

  6. Angiography-based C-arm CT for the assessment of extrahepatic shunting before radioembolization

    International Nuclear Information System (INIS)

    Heusner, Till Alexander; Hahn, S.; Forsting, M.; Antoch, G.; Hamami, M.E.; Poeppel, T.; Bockisch, A.; Ertle, J.; Hilgard, P.

    2010-01-01

    Purpose: to retrospectively assess the accuracy of angiography-based C-arm CT for the detection of extrahepatic shunting before SIRT. Materials and methods: 30 patients (mean age: 64 ± 12 years) with hypervascularized hepatic tumors underwent hepatic angiography, coil embolization of gastrointestinal collaterals and 99mTc-macroaggregated albumin (MAA) SPECT/CT before SIRT. Before MAA injection via a microcatheter from the intended treatment position, an angiography and angiography-based C-arm CT (XperCT trademark, Philips Healthcare) were acquired. Angiographies and XperCT trademark were performed from 48 microcatheter positions followed by MAA injections and MAA-SPECT/CT. MAA-SPECT/CT served as the reference standard for determining the accuracy of hepatic arteriography and C-arm CT for the detection of extrahepatic shunting. Results: MAA-SPECT/CT revealed extrahepatic shunting in 5 patients (17%). Hepatic arteriography yielded a true negative in 22 (73%), a false negative in 5 (17%), and an unclear result in 3 patients (10%). C-arm CT yielded a true positive in 3 (10%), true negative in 24 (80%), false positive in 1 (3%), and false negative in 2 patients (7%). The specificity and the NPV of hepatic arteriography for the detection of extrahepatic shunting were 88% and 81%, respectively. For C-arm CT the sensitivity, specificity, PPV, NPV, and accuracy for the detection of extrahepatic shunting were 60%, 96%, 75%, 92%, and 90%, respectively. Conclusion: C-arm CT offers additional information to angiography when assessing SIRT patients for extrahepatic shunting. More accurate detection of extrahepatic shunting may optimize the workflow in SIRT preparations by avoiding unnecessary repeat angiographies. (orig.)

  7. CT-guided percutaneous core needle biopsy in deep seated musculoskeletal lesions: a prospective study of 128 cases

    International Nuclear Information System (INIS)

    Puri, A.; Shingade, V.U.; Agarwal, M.G.; Anchan, C.; Juvekar, S.; Desai, S.; Jambhekar, N.A.

    2006-01-01

    Although large lesions of the limbs can easily be biopsied without image guidance, lesions in the spine, paraspinal area and pelvis are difficult to target, and benefit from CT guidance to improve the accuracy of targeting the lesion for biopsy purposes. A prospective study of CT-guided core needle biopsies for deep-seated musculoskeletal lesions was conducted at a referral cancer institute over a 4-year period with the aim of assessing the safety and efficacy of the procedure. From January 2000 to December 2003, 136 consecutive CT-guided biopsy sessions were undertaken for musculoskeletal lesions in 128 patients comprising 73 males and 55 females. The following data was recorded in all patients: demographic data, suspected clinicoradiological diagnosis, data related to core biopsy session (date, site, approach, total time required in minutes, number of cores, surgeon satisfaction with adequacy of cores), patient discomfort, complications, histopathology report and number of further sessions if material obtained during the first biopsy session was not confirmatory. The sample obtained during the biopsy session was considered inconclusive if, in the opinion of the pathologist, inadequate or non-representative tissue had been obtained. The diagnosis was considered inaccurate if the final histopathological diagnosis did not match with the biopsy diagnosis, or if subsequent clinicoradiological evaluation at follow up did not correlate with the biopsy diagnosis in those patients who were treated with modalities other than surgery. In 121 patients, a single session was sufficient to obtain representative material, whilst for six patients two sessions, and for one patient three sessions were necessary. The time taken for biopsy, including the pre-biopsy CT examination time, varied from 15 min to 60 min (median 30 min). For 110 bony lesions 116 sessions were required, and for 18 soft-tissue lesions 20 sessions were required. 108 biopsy sessions yielded a diagnosis, whilst

  8. The accuracy of CT and tumor markers in the detection of a recurrent ovarian carcinoma

    International Nuclear Information System (INIS)

    Wakabayashi, Yukari; Ishida, Jiro; Kotake, Fumio; Hirose, Masahiro; Kawana, Koji; Abe, Kimihiko; Amino, Saburo; Negishi, Yoshiyuki; Akiya, Kiyoshi

    1989-01-01

    Twenty-three patients previously diagnosed as having ovarian cancer were examined with both serum tumor markers (CA 125, CA 19-9, TPA, IAP, AFP) and a pelvic CT scan. The tumor markers predict the clinical outcome more accurately than the CT scan. Further, the tumor markers showed a clear correlation with the clinical course. But in one case, however, the tumor markers were seen to reduce below the normal level from chemotherapy, while the CT scan showed a tumor mass. Thus, both, a CT scan and tumor marker assays are felt to be indispensable for detecting the recurrence of an ovarian cancer. (author)

  9. FDG PET/CT diagnostic criteria may need adjustment based on MRI to estimate the presurgical risk of extrapelvic infiltration in patients with uterine endometrial cancer

    Energy Technology Data Exchange (ETDEWEB)

    Sudo, Satoko; Sakuragi, Noriaki [Hokkaido University Graduate School of Medicine, Department of Gynecology, Sapporo (Japan); Hattori, Naoya; Manabe, Osamu; Hirata, Kenji; Tamaki, Nagara [Hokkaido University Graduate School of Medicine, Department of Nuclear Medicine, Kitaku, Sapporo (Japan); Kato, Fumi; Mimura, Rie; Magota, Keiichi; Sugimori, Hiroyuki [Hokkaido University Graduate School of Medicine, Department of Diagnostic and Interventional Radiology, Sapporo (Japan)

    2015-04-01

    The staging of endometrial cancer requires surgery which carries the risk of morbidity. FDG PET/CT combined with anatomical imaging may reduce the number of unnecessary lymphadenectomies by demonstrating the risk of extrapelvic infiltration. The purpose of this study was to optimize FDG PET/CT diagnostic criteria for risk assessment in endometrial cancer after first-line risk triage with MRI. The study population comprised 37 patients who underwent curative surgery for the treatment of endometrial cancer. First, the risk of extrapelvic infiltration was triaged using MRI. Second, multiple glucose metabolic profiles of the primary lesion were assessed with FDG PET/CT, and these were correlated with the histopathological risk of extrapelvic infiltration including lymphovascular space invasion (LVSI) and high-grade malignancy (grades 2 and 3). The results of histological correlation were used to adjust FDG PET/CT diagnostic criteria. Presurgical assessment using MRI was positive for deep (>50 %) myometrial invasion in 17 patients. The optimal FDG PET/CT diagnostic criteria vary depending on the results of MRI. Specifically, SUVmax (≥16.0) was used to indicate LVSI risk with an overall diagnostic accuracy of 88.2 % in patients with MRI findings showing myometrial invasion. High-grade malignancy did not correlate with any of metabolic profiles in this patient group. In the remaining patients without myometrial invasion, lesion glycolysis (LG) or metabolic volume were better indicators of LVSI than SUVmax with the same diagnostic accuracy of 80.0 %. In addition, LG (≥26.9) predicted high-grade malignancy with an accuracy of 72.2 %. Using the optimized cut-off criteria for LVSI, glucose metabolic profiling of primary lesions correctly predicted lymph node metastasis with an accuracy of 73.0 %, which was comparable with the accuracy of visual assessment for lymph node metastasis using MRI and FDG PET/CT. FDG PET/CT diagnostic criteria may need adjustment based on the

  10. Comparison of the diagnostic value of FDG-PET/CT and axillary ultrasound for the detection of lymph node metastases in breast cancer patients

    International Nuclear Information System (INIS)

    Riegger, Carolin; Heusner, Till A.; Koeninger, Angela; Kimmig, Rainer; Hartung, Verena; Bockisch, Andreas; Otterbach, Friedrich; Forsting, Michael; Antoch, Gerald

    2012-01-01

    Background. FDG-PET/CT is increasingly being used for breast cancer staging. Its diagnostic accuracy in comparison to ultrasound as the standard non-invasive imaging modality for the evaluation of axillary lymph nodes has yet not been evaluated. Purpose. To retrospectively compare the diagnostic value of full-dose, intravenously contrast-enhanced FDG-PET/CT and ultrasound for the detection of lymph node metastases in breast cancer patients. Material and Methods. Ninety patients (one patient with a bilateral carcinoma) (89 women, one man; mean age, 55.5 ± 16.6 years) suffering from primary breast cancer underwent whole-body FDG-PET/CT and axillary ultrasound. The ipsilateral axillary fossa (n = 91) was evaluated for metastatic spread. The sensitivity, specificity, the positive predictive value (PPV), negative predictive value (NPV), and accuracy of both methods were calculated. The sensitivity and accuracy were statistically compared using the McNemar Test (P <0.05). Analyses were made on a patient basis. The number of patients with extra-axillary locoregional lymph node metastases exclusively detected by FDG-PET/CT was evaluated. For axillary lymph node metastases histopathology served as the reference standard. Results. The sensitivity, specificity, PPV, NPV, and accuracy of FDG-PET/CT for the detection of axillary lymph node metastases were 54%, 89%, 77%, 74%, and 75%, respectively. For ultrasound it was 38%, 78%, 54%, 65%, and 62%, respectively. FDG-PET/CT was significantly more accurate than ultrasound for the detection of axillary lymph node metastases (P = 0.019). There was no statistically significant difference between the sensitivity of both modalities (P = 0.0578). FDG-PET/CT detected extra-axillary locoregional lymph node metastases in seven patients (8%) that had not been detected by another imaging modality. Conclusion. Though more accurate compared to ultrasound for evaluating the axillary lymph node status FDG-PET/CT is only as sensitive as

  11. Sensitivity and accuracy of volumetry of pulmonary nodules on low-dose 16- and 64-row multi-detector CT: an anthropomorphic phantom study

    Energy Technology Data Exchange (ETDEWEB)

    Xie, Xueqian; Zhao, Yingru; Ooijen, Peter M.A. van; Vliegenthart, Rozemarijn [University of Groningen, University Medical Center Groningen, Department of Radiology, EB44, P.O. Box 30.001, Groningen (Netherlands); University of Groningen, University Medical Center Groningen, Center for Medical Imaging-North East Netherlands, Department of Radiology, Groningen (Netherlands); Snijder, Roland A.; Greuter, Marcel J.W. [University of Groningen, University Medical Center Groningen, Department of Radiology, EB44, P.O. Box 30.001, Groningen (Netherlands); Jong, Pim A. de [University Medical Center Utrecht, Department of Radiology, Utrecht (Netherlands); Oudkerk, Matthijs [University of Groningen, University Medical Center Groningen, Center for Medical Imaging-North East Netherlands, Department of Radiology, Groningen (Netherlands); Bock, Geertruida H. de [University of Groningen, University Medical Center Groningen, Department of Epidemiology, Groningen (Netherlands)

    2013-01-15

    To assess the sensitivity of detection and accuracy of volumetry by manual and semi-automated quantification of artificial pulmonary nodules in an anthropomorphic thoracic phantom on low-dose CT. Fifteen artificial spherical nodules (diameter 3, 5, 8, 10 and 12 mm; CT densities -800, -630 and +100 HU) were randomly placed inside an anthropomorphic thoracic phantom. The phantom was examined on 16- and 64-row multidetector CT with a low-dose protocol. Two independent blinded observers screened for pulmonary nodules. Nodule diameter was measured manually, and volume calculated. For solid nodules (+100 HU), diameter and volume were also evaluated by semi-automated software. Differences in observed volumes between the manual and semi-automated method were evaluated by a t-test. Sensitivity was 100 % for all nodules of >5 mm and larger, 60-80 % for solid and 0-20 % for non-solid 3-mm nodules. No false-positive nodules but high inter-observer reliability and inter-technique correlation were found. Volume was underestimated manually by 24.1 {+-} 14.0 % for nodules of any density, and 26.4 {+-} 15.5 % for solid nodules, compared with 7.6 {+-} 8.5 % (P < 0.01) semi-automatically. In an anthropomorphic phantom study, the sensitivity of detection is 100 % for nodules of >5 mm in diameter. Semi-automated volumetry yielded more accurate nodule volumes than manual measurements. (orig.)

  12. High-resolution CT of nontuberculous mycobacterium infection in adult CF patients: diagnostic accuracy

    International Nuclear Information System (INIS)

    McEvoy, Sinead; Lavelle, Lisa; Kilcoyne, Aoife; McCarthy, Colin; Dodd, Jonathan D.; DeJong, Pim A.; Loeve, Martine; Tiddens, Harm A.W.M.; McKone, Edward; Gallagher, Charles G.

    2012-01-01

    To determine the diagnostic accuracy of high-resolution computed tomography (HRCT) for the detection of nontuberculous mycobacterium infection (NTM) in adult cystic fibrosis (CF) patients. Twenty-seven CF patients with sputum-culture-proven NTM (NTM+) underwent HRCT. An age, gender and spirometrically matched group of 27 CF patients without NTM (NTM-) was included as controls. Images were randomly and blindly analysed by two readers in consensus and scored using a modified Bhalla scoring system. Significant differences were seen between NTM (+) and NTM (-) patients in the severity of the bronchiectasis subscore [45 % (1.8/4) vs. 35 % (1.4/4), P = 0.029], collapse/consolidation subscore [33 % (1.3/3) vs. 15 % (0.6/3)], tree-in-bud/centrilobular nodules subscore [43 % (1.7/3) vs. 25 % (1.0/3), P = 0.002] and the total CT score [56 % (18.4/33) vs. 46 % (15.2/33), P = 0.002]. Binary logistic regression revealed BMI, peribronchial thickening, collapse/consolidation and tree-in-bud/centrilobular nodules to be predictors of NTM status (R 2 = 0.43). Receiver-operator curve analysis of the regression model showed an area under the curve of 0.89, P < 0.0001. In adults with CF, seven or more bronchopulmonary segments showing tree-in-bud/centrilobular nodules on HRCT is highly suggestive of NTM colonisation. (orig.)

  13. High-resolution CT of nontuberculous mycobacterium infection in adult CF patients: diagnostic accuracy

    Energy Technology Data Exchange (ETDEWEB)

    McEvoy, Sinead; Lavelle, Lisa; Kilcoyne, Aoife; McCarthy, Colin; Dodd, Jonathan D. [St. Vincent' s University Hospital, Department of Radiology, Dublin (Ireland); DeJong, Pim A. [University Medical Center Utrecht, Department of Radiology, Utrecht (Netherlands); Loeve, Martine; Tiddens, Harm A.W.M. [Erasmus MC-Sophia Children' s Hospital, Department of Radiology, Department of Pediatric Pulmonology and Allergology, Rotterdam (Netherlands); McKone, Edward; Gallagher, Charles G. [St. Vincent' s University Hospital, Department of Respiratory Medicine and National Referral Centre for Adult Cystic Fibrosis, Dublin (Ireland)

    2012-12-15

    To determine the diagnostic accuracy of high-resolution computed tomography (HRCT) for the detection of nontuberculous mycobacterium infection (NTM) in adult cystic fibrosis (CF) patients. Twenty-seven CF patients with sputum-culture-proven NTM (NTM+) underwent HRCT. An age, gender and spirometrically matched group of 27 CF patients without NTM (NTM-) was included as controls. Images were randomly and blindly analysed by two readers in consensus and scored using a modified Bhalla scoring system. Significant differences were seen between NTM (+) and NTM (-) patients in the severity of the bronchiectasis subscore [45 % (1.8/4) vs. 35 % (1.4/4), P = 0.029], collapse/consolidation subscore [33 % (1.3/3) vs. 15 % (0.6/3)], tree-in-bud/centrilobular nodules subscore [43 % (1.7/3) vs. 25 % (1.0/3), P = 0.002] and the total CT score [56 % (18.4/33) vs. 46 % (15.2/33), P = 0.002]. Binary logistic regression revealed BMI, peribronchial thickening, collapse/consolidation and tree-in-bud/centrilobular nodules to be predictors of NTM status (R{sup 2} = 0.43). Receiver-operator curve analysis of the regression model showed an area under the curve of 0.89, P < 0.0001. In adults with CF, seven or more bronchopulmonary segments showing tree-in-bud/centrilobular nodules on HRCT is highly suggestive of NTM colonisation. (orig.)

  14. Prospective versus retrospective ECG gating for dual source CT of the coronary stent: Comparison of image quality, accuracy, and radiation dose

    Energy Technology Data Exchange (ETDEWEB)

    Zhao Lei, E-mail: zhaolei219@sohu.com [Beijing Anzhen Hospital of the Capital University of Medical Sciences (China); Zhang Zhaoqi; Fan Zhanming; Yang Lin; Du Jing [Beijing Anzhen Hospital of the Capital University of Medical Sciences (China)

    2011-03-15

    Objective: To compare image quality, diagnostic accuracy and radiation dose of prospective and retrospective electrocardiogram (ECG) gated dual source computed tomography (DSCT) for the evaluation of the coronary stent, using conventional coronary angiography (CA) as a standard reference. Design, setting and patients: Sixty patients (heart rates {<=}70 bpm) with previous stent implantation who were scheduled for CA were divided in two groups, receiving either prospective or retrospective ECG gated DSCT separately. Two reviewers scored coronary stent image quality and evaluated stent lumen. Results: There was no significant difference in image quality between the two groups. In the prospective group, there were 86.4% (51/59) stents with interpretable images, in the retrospective group, there were 87.5% (49/56) stents with interpretable images. Image quality was not influenced by age, body mass index or heart rate in either group, but heart rate variability had a weak impact on the image quality of the prospective group. Image noise was higher in the prospective group, but this difference reached statistical significance only by using a smooth kernel reconstruction. Per-stent based sensitivity, specificity, and positive and negative predictive value were 100%, 84.1%, 68.2%, and 100%, respectively, in the prospective CT angiography group and 94.4%, 86.8%, 77.3%, and 97.1%, respectively, in the retrospective CT angiography group. There was a significant difference in the effective radiation dose between the two groups, mean effective dose in the prospective and retrospective group was 2.2 {+-} 0.5 mSv (1.5-3.2 mSv) and 14.6 {+-} 3.3 mSv (10.0-20.4 mSv) (p < .001) respectively. Conclusions: Compared with retrospective CT angiography, prospective CT angiography has a similar performance in assessing coronary stent patency, but a lower effective dose in selected patients with regular heart rates {<=}70 bpm.

  15. CT diagnosis of rectal cancer

    International Nuclear Information System (INIS)

    Kanda, Hiroshi; Hachisuka, Kitao; Yamaguchi, Akihiro

    1986-01-01

    Preoperative diagnosis of the depth of invasion and lymph node metastasis of rectal cancer were studied using the findings of computed tomography (CT). Of one hundred and four cases operated on for rectal cancer over a period of 32 months, thirty five cases were examined by CT with the use of olive oil enema and contrast enhancement using a 60 % Conray drip infusion with reference to the histological findings. For direct invasion into the wall, the diagnoses by CT were coincident with microscopic findings in 75 % of cancers of the rectosigmoid, in 75 % of the upper rectum and in 84 % of the lower rectum. Of all cases, 28 (80 %) were diagnosed correctly. As to local lymph node metastasis, 74 % of all diagnoses by CT corresponded with the histological diagnosis. Moreover, seventeen cases were evaluated for lateral lymph node metastasis, and the diagnostic accuracy by CT was 88 %. In conclusion, preoperative CT evaluation of the extension into the rectal wall and lymph node metastasis in rectal cancer was considesed useful. (author)

  16. Can CT and MR Shape and Textural Features Differentiate Benign Versus Malignant Pleural Lesions?

    Science.gov (United States)

    Pena, Elena; Ojiaku, MacArinze; Inacio, Joao R; Gupta, Ashish; Macdonald, D Blair; Shabana, Wael; Seely, Jean M; Rybicki, Frank J; Dennie, Carole; Thornhill, Rebecca E

    2017-10-01

    The study aimed to identify a radiomic approach based on CT and or magnetic resonance (MR) features (shape and texture) that may help differentiate benign versus malignant pleural lesions, and to assess if the radiomic model may improve confidence and accuracy of radiologists with different subspecialty backgrounds. Twenty-nine patients with pleural lesions studied on both contrast-enhanced CT and MR imaging were reviewed retrospectively. Three texture and three shape features were extracted. Combinations of features were used to generate logistic regression models using histopathology as outcome. Two thoracic and two abdominal radiologists evaluated their degree of confidence in malignancy. Diagnostic accuracy of radiologists was determined using contingency tables. Cohen's kappa coefficient was used to assess inter-reader agreement. Using optimal threshold criteria, sensitivity, specificity, and accuracy of each feature and combination of features were obtained and compared to the accuracy and confidence of radiologists. The CT model that best discriminated malignant from benign lesions revealed an AUC CT  = 0.92 ± 0.05 (P textural and shape analysis may help distinguish malignant from benign lesions. A radiomics-based approach may increase diagnostic confidence of abdominal radiologists on CT and MR and may potentially improve radiologists' accuracy in the assessment of pleural lesions characterized by MR. Copyright © 2017 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

  17. Accuracy of renal volume assessment in children by three-dimensional sonography

    International Nuclear Information System (INIS)

    Fritz, G.A.; Riccabona, M.; Bohdal, G.; Quehenberger, F.

    2003-01-01

    Purpose: Prospective evaluation of the accuracy of three-dimensional ultrasound (3DUS) to assess the renal parenchymal volume. Materials and Methods: CT, MRI, 2DUS and 3DUS were used to measure the renal volume in 40 patients (range: neonate to 17 years; mean age: 8.95 years). The 3DUS was determined with a Voluson 730 (Kretztechnik, GE) or an external 3D-system (EchoTech, GE) using electromagnetic positioning sensors attached to conventional 2DUS-equipment. The 2DUS volume was calculated with the ellipsoid equation and the 3DUS volume computed with the system integrated software. For CT and MRI, planimetric analysis was used to determine the renal parenchymal volume, whereby the dilated collecting system of a hydronephrosis was subtracted to obtain the real renal parenchymal volume. The results of 2DUS and 3DUS were compared to the results of CT and MRI, and inter- and intraobserver variabilities were calculated. Results: In 74 of 77 kidneys, the 3DUS study was of diagnostic quality. The accuracy of the 3DUS volumes compared well to the CT and MRI volumes with a mean difference of -1.8 ± 4.6% versus a mean difference of -2.4 ± 15.4% for 2DUS. In normal kidneys, the accuracy was -2.6 ± 4.4% for 3DUS and -3.8 ± 14.7% for 2DUS. In hydronephrosis, the accuracy was +4.0 ± 5.9% and +9.6 ± 21.3%, respectively, indicating that 3DUS is more accurate than 2DUS, particularly in kidneys with a dilated collecting system. Inter- and intraobserver variabilities were ± 7.3% and ± 5.3%. Conclusion: For assessing the renal parenchymal volume in children, 3DUS is feasible and comparable to CT and MRI. (orig.) [de

  18. Role of 18F-FDG PET/CT in diagnosing peritoneal carcinomatosis in the restaging of patient with ovarian cancer as compared to contrast enhanced CT and tumor marker Ca-125.

    Science.gov (United States)

    Rubini, G; Altini, C; Notaristefano, A; Merenda, N; Rubini, D; Ianora, A A Stabile; Asabella, A Niccoli

    2014-01-01

    To investigate the role of whole-body fluorine-18-2-deoxy-2-fluoro-d-glucose positron emission tomography/computed tomography ((18)F-FDG PET/CT) in the identification of peritoneal carcinomatosis in patients with ovarian cancer (OC). Seventy-nine patients with histologically proven stages III-IV OC who underwent (18)F-FDG PET/CT were studied retrospectively. We considered group A as 51 patients who also underwent computed-tomography with contrast-enhancement (CECT), and group B as 35 patients who had also been tested for biomarker Ca-125. Sensitivity, specificity, accuracy, positive predictive values (PPV) and negative predictive values (NPV) of (18)F-FDG PET/CT as compared to CECT and to Ca-125 were evaluated. (18)F-FDG PET/CT' sensitivity, specificity, accuracy, PPV and NPV for all 79 patients were: 85%, 92.31%, 88.61%, 91.89% and 85.71%, respectively. (18)F-FDG PET/CT sensitivity in group A was 78.6%, while it was 53.6% for CECT. (18)F-FDG PET/CT specificity, calculated in the same group, was 91.3%, while that of CECT was 60.9% (statistically significant difference, McNemar 4, P=0.039). Accuracy was 84.3% and 56.9%, respectively. (18)F-FDG PET/CT' sensitivity in group B was 86.4%, while that of Ca-125 was 81.8% (no statistical difference, McNemar 0, P=1). (18)F-FDG PET/CT specificity in group B was 84.6% while that of Ca-125 was 38.5% (clear but not statistically significant difference, McNemar 3.12, P=0.070). Accuracy calculated in the same group was 85.7% for (18)F-FDG PET/CT and 65.7% for Ca-125. (18)F-FDG PET/CT is a useful diagnostic tool when peritoneal biopsy cannot be performed and it can better select those who are candidates for adjuvant chemotherapy. Copyright © 2013 Elsevier España, S.L. and SEMNIM. All rights reserved.

  19. A new method for measuring coronary artery diameters with CT spatial profile curves

    International Nuclear Information System (INIS)

    Shimamoto, Ryoichi; Suzuki, Jun-ichi; Yamazaki, Tadashi; Tsuji, Taeko; Ohmoto, Yuki; Morita, Toshihiro; Yamashita, Hiroshi; Honye, Junko; Nagai, Ryozo; Akahane, Masaaki; Ohtomo, Kuni

    2007-01-01

    Purpose: Coronary artery vascular edge recognition on computed tomography (CT) angiograms is influenced by window parameters. A noninvasive method for vascular edge recognition independent of window setting with use of multi-detector row CT was contrived and its feasibility and accuracy were estimated by intravascular ultrasound (IVUS). Methods: Multi-detector row CT was performed to obtain 29 CT spatial profile curves by setting a line cursor across short-axis coronary angiograms processed by multi-planar reconstruction. IVUS was also performed to determine the reference coronary diameter. IVUS diameter was fitted horizontally between two points on the upward and downward slopes of the profile curves and Hounsfield number was measured at the fitted level to test seven candidate indexes for definition of intravascular coronary diameter. The best index from the curves should show the best agreement with IVUS diameter. Results: Of the seven candidates the agreement was the best (agreement: 16 ± 11%) when the two ratios of Hounsfield number at the level of IVUS diameter over that at the peak on the profile curves were used with water and with fat as the background tissue. These edge definitions were achieved by cutting the horizontal distance by the curves at the level defined by the ratio of 0.41 for water background and 0.57 for fat background. Conclusions: Vascular edge recognition of the coronary artery with CT spatial profile curves was feasible and the contrived method could define the coronary diameter with reasonable agreement

  20. Regional PET/CT after water gastric inflation for evaluating loco-regional disease of gastric cancer

    International Nuclear Information System (INIS)

    Lee, Soo Jin; Lee, Won Woo; Yoon, Hai-Jeon; Lee, Ho-Young; Lee, Kyoung Ho; Kim, Young Hoon; Park, Do Joong; Kim, Hyung-Ho; So, Young

    2013-01-01

    Objective: We aimed to improve diagnostic accuracy of 18 F-fluoro-2-deoxyglucose (FDG) PET/CT for gastric cancer with water gastric inflation. Materials and methods: 44 gastric cancer patients (M:F = 30:14, age ± std = 62.1 ± 14.5y) were enrolled before surgery. Fifty minutes after injection of FDG (0.14 mCi/kg body weight), whole body PET/CT was performed first and then regional PET/CT over gastric area was obtained 80 min post FDG injection after water gastric inflation. Diagnostic accuracies for loco-regional lesions were compared between whole body and regional PET/CT. Results: 48 primary tumors (23 EGC and 25 AGC) and 348 LN stations (61 metastatic and 287 benign) in 44 patients were investigated. Primary tumor sensitivity of whole body PET/CT (50% = 24/48) was significantly improved by regional PET/CT (75% = 36/48, p < 0.005). Sensitivity of whole body PET/CT (24.6% = 15/61) for LN metastasis was also significantly improved by regional PET/CT (36.1% = 22/61, p < 0.01), whereas specificity of whole body PET/CT (99.3% = 285/287) was not compromised by regional PET/CT (98.3% = 282/287, p > 0.05). Higher primary tumor FDG uptake in regional PET/CT indicated shorter progress-free survival (p = 0.0003). Conclusion: Diagnostic accuracy of whole body PET/CT for loco-regional disease of gastric cancer could be significantly improved by regional PET/CT after water gastric inflation and prognosis could be effectively predicted by primary tumor FDG uptake in regional PET/CT

  1. Comparison of SPECT/CT, MRI and CT in diagnosis of skull base bone invasion in nasopharyngeal carcinoma.

    Science.gov (United States)

    Zhang, Shu-xu; Han, Peng-hui; Zhang, Guo-qian; Wang, Rui-hao; Ge, Yong-bin; Ren, Zhi-gang; Li, Jian-sheng; Fu, Wen-hai

    2014-01-01

    Early detection of skull base invasion in nasopharyngeal carcinoma (NPC) is crucial for correct staging, assessing treatment response and contouring the tumor target in radiotherapy planning, as well as improving the patient's prognosis. To compare the diagnostic efficacy of single photon emission computed tomography/computed tomography (SPECT/CT) imaging, magnetic resonance imaging (MRI) and computed tomography (CT) for the detection of skull base invasion in NPC. Sixty untreated patients with histologically proven NPC underwent SPECT/CT imaging, contrast-enhanced MRI and CT. Of the 60 patients, 30 had skull base invasion confirmed by the final results of contrast-enhanced MRI, CT and six-month follow-up imaging (MRI and CT). The diagnostic efficacy of the three imaging modalities in detecting skull base invasion was evaluated. The rates of positive findings of skull base invasion for SPECT/CT, MRI and CT were 53.3%, 48.3% and 33.3%, respectively. The sensitivity, specificity and accuracy were 93.3%, 86.7% and 90.0% for SPECT/CT fusion imaging, 96.7%, 100.0% and 98.3% for contrast-enhanced MRI, and 66.7%, 100.0% and 83.3% for contrast-enhanced CT. MRI showed the best performance for the diagnosis of skull base invasion in nasopharyngeal carcinoma, followed closely by SPECT/CT. SPECT/CT had poorer specificity than that of both MRI and CT, while CT had the lowest sensitivity.

  2. Diagnostic accuracy of fluorine-18-fluorodeoxyglucose positron emission tomography in gallbladder cancer: A meta-analysis.

    Science.gov (United States)

    Annunziata, Salvatore; Pizzuto, Daniele Antonio; Caldarella, Carmelo; Galiandro, Federica; Sadeghi, Ramin; Treglia, Giorgio

    2015-10-28

    To meta-analyze published data about the diagnostic accuracy of fluorine-18-fluorodeoxyglucose ((18)F-FDG) positron emission tomography (PET) and PET/computed tomography (PET/CT) in the evaluation of primary tumor in patients with gallbladder cancer (GBCa). A comprehensive literature search of studies published through 30(th) June 2014 regarding the role of (18)F-FDG PET and PET/CT in the evaluation of primary gallbladder cancer (GBCa) was performed. All retrieved studies were reviewed. Pooled sensitivity and specificity of (18)F-FDG PET or PET/CT in the evaluation of primary GBCa were calculated. The area under the summary receiving operator characteristics curve (AUC) was calculated to measure the accuracy of these methods. Sub-analyses considering the device used (PET vs PET/CT) were carried out. Twenty-one studies comprising 495 patients who underwent (18)F-FDG PET or PET/CT for suspicious GBCa were selected for the systematic review. The meta-analysis of 13 selected studies provided the following results: sensitivity 87% (95%CI: 82%-92%), specificity 78% (95%CI: 68%-86%). The AUC was 0.88. Improvement of sensitivity and specificity was observed when PET/CT was used. (18)F-FDG-PET and PET/CT demonstrated to be useful diagnostic imaging methods in the assessment of primary tumor in GBCa patients, nevertheless possible sources of false-negative and false-positive results should be kept in mind. PET/CT seems to have a better diagnostic accuracy than PET alone in this setting.

  3. Fast and accurate registration of cranial CT images with A-mode ultrasound.

    Science.gov (United States)

    Fieten, Lorenz; Schmieder, Kirsten; Engelhardt, Martin; Pasalic, Lamija; Radermacher, Klaus; Heger, Stefan

    2009-05-01

    Within the CRANIO project, a navigation module based on preoperative computed tomography (CT) data was developed for Computer and Robot Assisted Neurosurgery. The approach followed for non-invasive user-interactive registration of cranial CT images with the physical operating space consists of surface-based registration following pre-registration based on anatomical landmarks. Surface-based registration relies on bone surface points digitized transcutaneously by means of an optically tracked A-mode ultrasound (US) probe. As probe alignment and thus bone surface point digitization may be time-consuming, we investigated how to obtain high registration accuracy despite inaccurate pre-registration and a limited number of digitized bone surface points. Furthermore, we aimed at efficient man-machine-interaction during the probe alignment process. Finally, we addressed the problem of registration plausibility estimation in our approach. We modified the Iterative Closest Point (ICP) algorithm, presented by Besl and McKay and frequently used for surface-based registration, such that it can escape from local minima of the cost function to be iteratively minimized. The random-based ICP (R-ICP) we developed is less influenced by the quality of the pre-registration as it can escape from local minima close to the starting point for iterative optimization in the 6D domain of rigid transformations. The R-ICP is also better suited to approximate the global minimum as it can escape from local minima in the vicinity of the global minimum, too. Furthermore, we developed both CT-less and CT-based probe alignment tools along with appropriate man-machine strategies for a more time-efficient palpation process. To improve registration reliability, we developed a simple plausibility test based on data readily available after registration. In a cadaver study, where we evaluated the R-ICP algorithm, the probe alignment tools, and the plausibility test, the R-ICP algorithm consistently

  4. Accuracy of DXA scanning of the thoracic spine: cadaveric studies comparing BMC, areal BMD and geometric estimates of volumetric BMD against ash weight and CT measures of bone volume.

    Science.gov (United States)

    Sran, Meena M; Khan, Karim M; Keiver, Kathy; Chew, Jason B; McKay, Heather A; Oxland, Thomas R

    2005-12-01

    Biomechanical studies of the thoracic spine often scan cadaveric segments by dual energy X-ray absorptiometry (DXA) to obtain measures of bone mass. Only one study has reported the accuracy of lateral scans of thoracic vertebral bodies. The accuracy of DXA scans of thoracic spine segments and of anterior-posterior (AP) thoracic scans has not been investigated. We have examined the accuracy of AP and lateral thoracic DXA scans by comparison with ash weight, the gold-standard for measuring bone mineral content (BMC). We have also compared three methods of estimating volumetric bone mineral density (vBMD) with a novel standard-ash weight (g)/bone volume (cm3) as measured by computed tomography (CT). Twelve T5-T8 spine segments were scanned with DXA (AP and lateral) and CT. The T6 vertebrae were excised, the posterior elements removed and then the vertebral bodies were ashed in a muffle furnace. We proposed a new method of estimating vBMD and compared it with two previously published methods. BMC values from lateral DXA scans displayed the strongest correlation with ash weight (r=0.99) and were on average 12.8% higher (pBMC (AP or lateral) was more strongly correlated with ash weight than areal bone mineral density (aBMD; AP: r=0.54, or lateral: r=0.71) or estimated vBMD. Estimates of vBMD with either of the three methods were strongly and similarly correlated with volumetric BMD calculated by dividing ash weight by CT-derived volume. These data suggest that readily available DXA scanning is an appropriate surrogate measure for thoracic spine bone mineral and that the lateral scan might be the scan method of choice.

  5. CT-Urography; Urografia CT

    Energy Technology Data Exchange (ETDEWEB)

    Dalla Palma, Ludovico; Grotto, Maurizio [Trieste Univ., Trieste (Italy). Dipartimento di scienze cliniche, morfologiche e tecnologiche, UCO di radiologia; Morra, Aldo [CRO, Aviano (Italy). Reparto di radiologia

    2005-09-15

    In this paper we present an overview of CT-Urography. With the advent of multislice CT scanners and the evolution of image processing methods this technique now affords optimal urographic images comparable to those obtained with conventional techniques. We describe the acquisition techniques and protocols used by the various authors. Effective radiation dose has conditioned the use of CT-Urography so that the tendency today is to reduce the number of scans by performing, after the non enhanced scan, a single contrast-enhanced scan comprising both the nephrographic and urographic phase. With the use of multislice CT the quality of the urogram improves with the number of slices. We illustrate a variety of processing techniques, multiplanar reconstruction (MPR), maximum (MIP) and average intensity projection (AIP) and volume rendering (VR) and present a series of upper urinary tract tumours testifying to the superiority of the AIP technique over MIP. We then review the results of comparative studies of CT-Urography with conventional urography in upper urinary tract diagnostics. Finally, we describe the advantages and limitations of CT-Urography. [Italian] Gli Autori presentano una panoramica sulla Urografia TC (Uro TC). L'avvento della TC multistrato e l'evoluzione delle tecniche di elaborazione consentono di ottenere dei quadri urografici ottimali comparativi con quelli convenzionali. Vengono ricordate le varie tecniche di acquisizione e i protocolli usati dai vari Autori. La dose effettiva di radiazioni ha rappresentato uno dei fattori condizionanti per cui oggi prevale l'orientamento di ridurre il numero di scansioni, considerando dopo la scansione senza mezzo di contrasto un'unica scansione contrasto grafica comprendente sia la fase nefrografica che quella urografica. Con l'uso della TC multistrato la qualita dell'urogramma migliora con l'aumento del numero degli strati. Vengono descritte le varie tecniche di elaborazione, la

  6. A combination-weighted Feldkamp-based reconstruction algorithm for cone-beam CT

    International Nuclear Information System (INIS)

    Mori, Shinichiro; Endo, Masahiro; Komatsu, Shuhei; Kandatsu, Susumu; Yashiro, Tomoyasu; Baba, Masayuki

    2006-01-01

    The combination-weighted Feldkamp algorithm (CW-FDK) was developed and tested in a phantom in order to reduce cone-beam artefacts and enhance cranio-caudal reconstruction coverage in an attempt to improve image quality when utilizing cone-beam computed tomography (CBCT). Using a 256-slice cone-beam CT (256CBCT), image quality (CT-number uniformity and geometrical accuracy) was quantitatively evaluated in phantom and clinical studies, and the results were compared to those obtained with the original Feldkamp algorithm. A clinical study was done in lung cancer patients under breath holding and free breathing. Image quality for the original Feldkamp algorithm is degraded at the edge of the scan region due to the missing volume, commensurate with the cranio-caudal distance between the reconstruction and central planes. The CW-FDK extended the reconstruction coverage to equal the scan coverage and improved reconstruction accuracy, unaffected by the cranio-caudal distance. The extended reconstruction coverage with good image quality provided by the CW-FDK will be clinically investigated for improving diagnostic and radiotherapy applications. In addition, this algorithm can also be adapted for use in relatively wide cone-angle CBCT such as with a flat-panel detector CBCT

  7. High impact of FDG-PET/CT in diagnostic strategies for ovarian cancer

    International Nuclear Information System (INIS)

    Zytoon, Ashraf Anas; Murakami, Koji; Eid, Hazem; El-Gammal, Mahmoud

    2013-01-01

    Background: Ovarian cancer has the highest mortality of all gynecologic malignancies. FDG-PET/CT was proven to be accurate for identification of primary ovarian tumors, regional lymph nodes, and distant metastases. Purpose: To evaluate ovarian masses at FDG-PET/CT in correlation with histopathologic findings. Material and Methods: Ninety-eight patients underwent whole body FDG-PET/CT examination. Eighty-six patients with primary ovarian cancer and 12 patients with metastatic disease to the ovaries were included. Results: PET/CT imaging was true-positive in 87/94 patients with malignant tumors. In 4/4 patients with benign tumors, PET/CT results were true-negative, with sensitivity of 92.6%, specificity 100%, total test accuracy 92.9%. Fifty-seven patients were diagnosed as stage IV ovarian cancer with distant metastasis. Conclusion: The anatomical/functional examination by FDG-PET/CT was proven to be valuable in increasing the diagnostic accuracy that can help improve patient management

  8. Planning and clinical studies of a commercial orthopedic metal artifact reduction tool for CT simulations for head-and-neck radiotherapy

    International Nuclear Information System (INIS)

    Kon, Hyuck Jun; Ye, Sung Joon; Kim, Jung In; Park, Jong Min; Lee, Jae Gi; Heo, Tae Min; Kim Kyung Su; Chun, Young Mi; Callahan, Zachariah

    2013-01-01

    In computed tomography (CT) images, the presence of high Z materials induces typical streak artifacts, called metal artifacts which can pervert CT Hounsfield numbers in the reconstructed images. These artifact-induced distortion of CT images can impact on the dose calculation based on the CT images. In the radiation therapy of Head-and-Neck cancer because of the concave-shaped target volumes, the complex anatomy, a lot of sensitive normal tissues and air cavity structures, it is important to get accurate CT images for dose calculation. But dental implant is common for H and N patients so that it is hard to get undistorted CT images. Moreover because dental implants are generally with the air cavity like oral cavity and nasal cavity in the same CT slice, they can make lots of distortion. In this study, we focused on evaluating the distortion on air cavity by the metal artifact and the effectiveness of the commercial orthopedic metal artifact reduction function (O-MAR) about the metal artifacts induced by the dental implant. The O-MAR algorithm increases the accuracy of CT Hounsfield numbers and reducing noises. Thus, it can contribute to the entire radiation treatment planning process, especially for contouring/segmentation. Although there was no significant difference in dose distributions for most cases, the O-MAR correction was shown to have an impact on high dose regions in air cavity

  9. Planning and clinical studies of a commercial orthopedic metal artifact reduction tool for CT simulations for head-and-neck radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Kon, Hyuck Jun; Ye, Sung Joon [Interdisplinary Program in Radiation Applied Life Science, Seoul National University Graduate School, Seoul (Korea, Republic of); Kim, Jung In; Park, Jong Min; Lee, Jae Gi; Heo, Tae Min [Institute of Radiation Medicine, Seoul National University Medical Research Center, Seoul (Korea, Republic of); Kim Kyung Su [Dept. of Radiation Oncology, Seoul National University College of Medicine, Seoul (Korea, Republic of); Chun, Young Mi [Philips Healthcare Korea, Seoul (Korea, Republic of); Callahan, Zachariah [Program in Biomedical Radiation Sciences, Dept. of Transdisciplinary Studies, Graduate School of Convergence Science and Technology, Seoul National University, Seoul (Korea, Republic of)

    2013-11-15

    In computed tomography (CT) images, the presence of high Z materials induces typical streak artifacts, called metal artifacts which can pervert CT Hounsfield numbers in the reconstructed images. These artifact-induced distortion of CT images can impact on the dose calculation based on the CT images. In the radiation therapy of Head-and-Neck cancer because of the concave-shaped target volumes, the complex anatomy, a lot of sensitive normal tissues and air cavity structures, it is important to get accurate CT images for dose calculation. But dental implant is common for H and N patients so that it is hard to get undistorted CT images. Moreover because dental implants are generally with the air cavity like oral cavity and nasal cavity in the same CT slice, they can make lots of distortion. In this study, we focused on evaluating the distortion on air cavity by the metal artifact and the effectiveness of the commercial orthopedic metal artifact reduction function (O-MAR) about the metal artifacts induced by the dental implant. The O-MAR algorithm increases the accuracy of CT Hounsfield numbers and reducing noises. Thus, it can contribute to the entire radiation treatment planning process, especially for contouring/segmentation. Although there was no significant difference in dose distributions for most cases, the O-MAR correction was shown to have an impact on high dose regions in air cavity.

  10. Rectosigmoid endometriosis: Comparison between CT water enema and video laparoscopy

    International Nuclear Information System (INIS)

    Stabile Ianora, A.A.; Moschetta, M.; Lorusso, F.; Lattarulo, S.; Telegrafo, M.; Rella, L.; Scardapane, A.

    2013-01-01

    Aim: To evaluate the accuracy of water enema computed tomography (CT) for predicting the location of endometriosis in patients with contraindications to magnetic resonance imaging (MRI), focusing on rectosigmoid lesions and having laparoscopic and histological data as the reference standard. Materials and methods: Thirty-three women (mean age 33.4 ± 3.1 years) suspected of having deep pelvic endometriosis underwent 64-row CT and video laparoscopy within 4 weeks. Two radiologists blinded to the clinical data evaluated the CT images obtained after colonic retrograde distension using water as the contrast medium, and a comparison with laparoscopic and histological findings was performed. CT sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and diagnostic accuracy were calculated. The radiation dose to patients was estimated. Cohen's weighted kappa (κ) test was used to evaluate the interobserver agreement. Results: In 23 out of 33 patients (69%) intestinal implants were found at surgery and pathological examinations. CT confirmed the diagnosis of rectosigmoid endometriosis in 20 out of 23 implants. Three nodules located on the proximal sigmoid colon (two serosal lesions and one infiltrating the muscularis layer) with a diameter of less than 1 cm were not diagnosed. CT sensitivity, specificity, PPV, NPV, and accuracy values were 87, 100, 100, 77, and 91%, respectively. The mean effective dose estimate was 6.30 ± 1.7 mSv. Almost perfect agreement between the two readers was found (k = 0.84). Conclusion: Water enema CT can play a role in the diagnosis of bowel endometriosis and represents another accurate potential tool for video laparoscopic approaches, especially in patients for whom MRI is contraindicated

  11. Diagnostic performance of a Lattice Boltzmann-based method for CT-based fractional flow reserve.

    Science.gov (United States)

    Giannopoulos, Andreas A; Tang, Anji; Ge, Yin; Cheezum, Michael K; Steigner, Michael L; Fujimoto, Shinichiro; Kumamaru, Kanako K; Chiappino, Dante; Della Latta, Daniele; Berti, Sergio; Chiappino, Sara; Rybicki, Frank J; Melchionna, Simone; Mitsouras, Dimitrios

    2018-02-20

    Fractional flow reserve (FFR) estimated from coronary computed tomography angiography (CT-FFR) offers non-invasive detection of lesion-specific ischaemia. We aimed to develop and validate a fast CT-FFR algorithm utilising the Lattice Boltzmann method for blood flow simulation (LBM CT-FFR). Sixty-four patients with clinically indicated CTA and invasive FFR measurement from three institutions were retrospectively analysed. CT-FFR was performed using an onsite tool interfacing with a commercial Lattice Boltzmann fluid dynamics cloud-based platform. Diagnostic accuracy of LBM CT-FFR ≤0.8 and percent diameter stenosis >50% by CTA to detect invasive FFR ≤0.8 were compared using area under the receiver operating characteristic curve (AUC). Sixty patients successfully underwent LBM CT-FFR analysis; 29 of 73 lesions in 69 vessels had invasive FFR ≤0.8. Total time to perform LBM CT-FFR was 40±10 min. Compared to invasive FFR, LBM CT-FFR had good correlation (r=0.64), small bias (0.009) and good limits of agreement (-0.223 to 0.206). The AUC of LBM CT-FFR (AUC=0.894, 95% confidence interval [CI]: 0.792-0.996) was significantly higher than CTA (AUC=0.685, 95% CI: 0.576-0.794) to detect FFR ≤0.8 (p=0.0021). Per-lesion specificity, sensitivity, and accuracy of LBM CT-FFR were 97.7%, 79.3%, and 90.4%, respectively. LBM CT-FFR has very good diagnostic accuracy to detect lesion-specific ischaemia (FFR ≤0.8) and can be performed in less than one hour.

  12. PET/CT imaging in head and neck tumors

    International Nuclear Information System (INIS)

    Roedel, R.; Palmedo, H.; Reichmann, K.; Reinhardt, M.J.; Biersack, H.J.; Straehler-Pohl, H.J.; Jaeger, U.

    2004-01-01

    To evaluate the usefulness of combined PET/CT examinations for detection of malignant tumors and their metastases in head and neck oncology. 51 patients received whole body scans on a dual modality PET/CT system. CT was performed without i.v. contrast. The results were compared concerning the diagnostic impact of native CT scan on FDG-PET images and the additional value of fused imaging. From 153 lesions were 97 classified as malignant on CT and 136 on FDG/PET images, as suspicious for malignancy in 33 on CT and 7 on FDG-PET and as benign in 23 on CT and 10 on FDG-PET. With combined PET/CT all primary and recurrent tumors could be found, the detection rate in patients with unknown primary tumors was 45%. Compared to PET or CT alone the sensitivity, specifity and accuracy could be significantly improved by means of combined PET/CT. Fused PET/CT imaging with [F18]-FDG and native CT-scanning enables accurate diagnosis in 93% of lesions and 90% of patients with head and neck oncology. (orig.) [de

  13. Assessment of the diagnostic accuracy of {sup 18}F-FDG PET/CT in prosthetic infective endocarditis and cardiac implantable electronic device infection: comparison of different interpretation criteria

    Energy Technology Data Exchange (ETDEWEB)

    Jimenez-Ballve, Ana; Jesus Perez-Castejon, Maria; Carreras-Delgado, Jose L. [Clinico San Carlos University Hospital, San Carlos Health Research Institute (IdISSC), Complutense University of Madrid, Department of Nuclear Medicine, Madrid (Spain); Delgado-Bolton, Roberto C. [Clinico San Carlos University Hospital, San Carlos Health Research Institute (IdISSC), Complutense University of Madrid, Department of Nuclear Medicine, Madrid (Spain); San Pedro Hospital and Centre for Biomedical Research of La Rioja (CIBIR), University of La Rioja, Department of Diagnostic Imaging (Radiology) and Nuclear Medicine, La Rioja (Spain); Sanchez-Enrique, Cristina; Vilacosta, Isidre; Vivas, David; Olmos, Carmen [Clinico San Carlos University Hospital, Department of Cardiology, Madrid (Spain); Ferrer, Manuel E.F. [Clinico San Carlos University Hospital, Research Unit, Madrid (Spain)

    2016-12-15

    The diagnosis of prosthetic valve (PV) infective endocarditis (IE) and infection of cardiac implantable electronic devices (CIEDs) remains challenging. The aim of this study was to assess the usefulness of {sup 18}F-FDG PET/CT in these patients and analyse the interpretation criteria. We included 41 patients suspected of having IE by the Duke criteria who underwent {sup 18}F-FDG PET/CT. The criteria applied for classifying the findings as positive/negative for IE were: (a) visual analysis of only PET images with attenuation-correction (AC PET images); (b) visual analysis of both AC PET images and PET images without AC (NAC PET images); (c) qualitative analysis of NAC PET images; and (d) semiquantitative analysis of AC PET images. {sup 18}F-FDG PET/CT was considered positive for IE independently of the intensity and distribution of FDG uptake. The gold standard was the Duke pathological criteria (if tissue was available) or the decision of an endocarditis expert team after a minimum 4 months follow-up. We studied 62 areas with suspicion of IE, 28 areas (45 %) showing definite IE and 34 (55 %) showing possible IE. Visual analysis of only AC PET images showed poor diagnostic accuracy (sensitivity 20 %, specificity 57 %). Visual analysis of both AC PET and NAC PET images showed excellent sensitivity (100 %) and intermediate specificity (73 %), focal uptake being more frequently associated with IE. The accuracy of qualitative analysis of NAC PET images depended on the threshold: the maximum sensitivity, specificity and accuracy achieved were 88 %, 80 %, 84 %, respectively. In the semiquantitative analysis of AC PET images, SUVmax was higher in areas of confirmed IE than in those without IE (∇SUVmax 2.2, p < 0.001). When FDG uptake was twice that in the liver, IE was always confirmed, and SUVmax 5.5 was the optimal threshold for IE diagnosis using ROC curve analysis (area under the curve 0.71). The value of {sup 18}F-FDG PET/CT in the diagnosis of suspected IE of PVs

  14. Assessment of the diagnostic accuracy of "1"8F-FDG PET/CT in prosthetic infective endocarditis and cardiac implantable electronic device infection: comparison of different interpretation criteria

    International Nuclear Information System (INIS)

    Jimenez-Ballve, Ana; Jesus Perez-Castejon, Maria; Carreras-Delgado, Jose L.; Delgado-Bolton, Roberto C.; Sanchez-Enrique, Cristina; Vilacosta, Isidre; Vivas, David; Olmos, Carmen; Ferrer, Manuel E.F.

    2016-01-01

    The diagnosis of prosthetic valve (PV) infective endocarditis (IE) and infection of cardiac implantable electronic devices (CIEDs) remains challenging. The aim of this study was to assess the usefulness of "1"8F-FDG PET/CT in these patients and analyse the interpretation criteria. We included 41 patients suspected of having IE by the Duke criteria who underwent "1"8F-FDG PET/CT. The criteria applied for classifying the findings as positive/negative for IE were: (a) visual analysis of only PET images with attenuation-correction (AC PET images); (b) visual analysis of both AC PET images and PET images without AC (NAC PET images); (c) qualitative analysis of NAC PET images; and (d) semiquantitative analysis of AC PET images. "1"8F-FDG PET/CT was considered positive for IE independently of the intensity and distribution of FDG uptake. The gold standard was the Duke pathological criteria (if tissue was available) or the decision of an endocarditis expert team after a minimum 4 months follow-up. We studied 62 areas with suspicion of IE, 28 areas (45 %) showing definite IE and 34 (55 %) showing possible IE. Visual analysis of only AC PET images showed poor diagnostic accuracy (sensitivity 20 %, specificity 57 %). Visual analysis of both AC PET and NAC PET images showed excellent sensitivity (100 %) and intermediate specificity (73 %), focal uptake being more frequently associated with IE. The accuracy of qualitative analysis of NAC PET images depended on the threshold: the maximum sensitivity, specificity and accuracy achieved were 88 %, 80 %, 84 %, respectively. In the semiquantitative analysis of AC PET images, SUVmax was higher in areas of confirmed IE than in those without IE (∇SUVmax 2.2, p < 0.001). When FDG uptake was twice that in the liver, IE was always confirmed, and SUVmax 5.5 was the optimal threshold for IE diagnosis using ROC curve analysis (area under the curve 0.71). The value of "1"8F-FDG PET/CT in the diagnosis of suspected IE of PVs and CIEDs is

  15. Frameless image registration of X-ray CT and SPECT by volume matching

    International Nuclear Information System (INIS)

    Tanaka, Yuko; Kihara, Tomohiko; Yui, Nobuharu; Kinoshita, Fujimi; Kamimura, Yoshitsugu; Yamada, Yoshifumi.

    1998-01-01

    Image registration of functional (SPECT) and morphological (X-ray CT/MRI) images is studied in order to improve the accuracy and the quantity of the image diagnosis. We have developed a new frameless registration method of X-ray CT and SPECT image using transmission CT image acquired for absorption correction of SPECT images. This is the automated registration method and calculates the transformation matrix between the two coordinate systems of image data by the optimization method. This registration method is based on the similar physical property of X-ray CT and transmission CT image. The three-dimensional overlap of the bone region is used for image matching. We verified by a phantom test that it can provide a good result of within two millimeters error. We also evaluated visually the accuracy of the registration method by the application study of SPECT, X-ray CT, and transmission CT head images. This method can be carried out accurately without any frames. We expect this registration method becomes an efficient tool to improve image diagnosis and medical treatment. (author)

  16. Value of noncontrast spiral CT for suspected acute appendicitis

    International Nuclear Information System (INIS)

    Choi, Pil Yeob; Lee, Sang Wook; Kwon, Jae Soo; Sung, Young Soon; Rho, Myoung Ho; Chang, Jeong A.

    1998-01-01

    To assess the diagnostic accuracy and clinical efficacy of noncontrast spiral CT in patients with suspected acute appendicitis. Over a six-month period, 100 patients with suspected acute appendicitis were prospectively evaluated with noncontrast spiral CT. All scans were obtained from the lower body of L3 to the symphysis pubis, with 5mm or 10mm collimation and pitch of 1 or 1.5, and without intravenous or oral contrast material. Diagnosis was established by means of surgical or clinical follow-up. Prospective diagnosis based on CT findings was compared with surgical results and clinical follow-up. Acute appendicitis was confirmed in 47 of 100 patients. On the basis of the Ct findings, SI patients were prospectively interpreted as positive for appendicitis, but in six the diagnosis was false-positive. Two of the 47 with acute appendicitis were prospectively interpreted as normal. The preoperative diagnosis of acute appendicitis was, thus, 45 true-positive, 47 true-negative, six false-positive and two false-negative yielding a sensitivity of 96%, a specificity of 89%, an accuracy of 92%, a positive predictive value of 88%, and a negative predictive value of 96%. Using CT, an alternative diagnosis was established in 14 patients. Noncontrast spiral CT is a useful technique for diagnosing acute appendicitis. =20

  17. PET/CT in cancer: moderate sample sizes may suffice to justify replacement of a regional gold standard

    DEFF Research Database (Denmark)

    Gerke, Oke; Poulsen, Mads Hvid; Bouchelouche, Kirsten

    2009-01-01

    PURPOSE: For certain cancer indications, the current patient evaluation strategy is a perfect but locally restricted gold standard procedure. If positron emission tomography/computed tomography (PET/CT) can be shown to be reliable within the gold standard region and if it can be argued that PET...... of metastasized prostate cancer. RESULTS: An added value in accuracy of PET/CT in adjacent areas can outweigh a downsized target level of accuracy in the gold standard region, justifying smaller sample sizes. CONCLUSIONS: If PET/CT provides an accuracy benefit in adjacent regions, then sample sizes can be reduced....../CT also performs well in adjacent areas, then sample sizes in accuracy studies can be reduced. PROCEDURES: Traditional standard power calculations for demonstrating sensitivities of both 80% and 90% are shown. The argument is then described in general terms and demonstrated by an ongoing study...

  18. Three-dimensional display improves observer speed and accuracy

    International Nuclear Information System (INIS)

    Nelson, J.A.; Rowberg, A.H.; Kuyper, S.; Choi, H.S.

    1989-01-01

    In an effort to evaluate the potential cost-effectiveness of three-dimensional (3D) display equipment, we compared the speed and accuracy of experienced radiologists identifying in sliced uppercase letters from CT scans with 2D and pseudo-3D display. CT scans of six capital letters were obtained and printed as a 2D display or as a synthesized pseudo-3D display (Pixar). Six observes performed a timed identification task. Radiologists read the 3D display an average of 16 times faster than the 2D, and the average error rate of 2/6 (± 0.6/6) for 2D interpretations was totally eliminated. This degree of improvement in speed and accuracy suggests that the expense of 3D display may be cost-effective in a defined clinical setting

  19. Hypervascular hepatocellular carcinomas: detection with gadoxetate disodium-enhanced MR imaging and multiphasic multidetector CT

    Energy Technology Data Exchange (ETDEWEB)

    Onishi, Hiromitsu; Kim, Tonsok; Hori, Masatoshi; Nakaya, Yasuhiro; Tsuboyama, Takahiro; Nakamoto, Atsushi; Tatsumi, Mitsuaki; Tomiyama, Noriyuki [Osaka University Graduate School of Medicine, Department of Radiology, Suita, Osaka (Japan); Imai, Yasuharu [Ikeda Municipal Hospital, Department of Gastroenterology, Ikeda, Osaka (Japan); Nagano, Hiroaki [Osaka University Graduate School of Medicine, Department of Surgery, Suita, Osaka (Japan); Kumano, Seishi; Okada, Masahiro; Murakami, Takamichi [Kinki University School of Medicine, Department of Radiology, Osakasayama, Osaka (Japan); Takamura, Manabu [Ikeda Municipal Hospital, Department of Radiology, Ikeda, Osaka (Japan); Wakasa, Kenichi [Osaka City University Graduate School of Medicine, Department of Diagnostic Pathology, Osaka, Osaka (Japan)

    2012-04-15

    To retrospectively compare the accuracy of detection of hypervascular hepatocellular carcinoma (HCC) by multiphasic multidetector CT and by gadoxetate disodium-enhanced MR imaging. After ethical approval, we analysed a total of 73 hypervascular HCC lesions from 31 patients suspected of having HCC, who underwent both gadoxetate disodium-enhanced MR imaging and multiphasic multidetector CT. Five blinded observers independently reviewed CT images, as well as dynamic MR images alone and combined with hepatobiliary phase MR images. Diagnostic accuracy (Az values), sensitivities and positive predictive values were compared by using the Scheffe post hoc test. The mean Az value for dynamic and hepatobiliary phase MR combined (0.81) or dynamic MR images alone (0.78) was significantly higher than that for CT images (0.67, P < 0.001, 0.005, respectively). The mean sensitivity of the combined MR images (0.67) was significantly higher than that of dynamic MR alone (0.52, P < 0.05) or CT images (0.44, P < 0.05). The mean positive predictive values were 0.96, 0.95 and 0.94, for CT, dynamic MR alone and combined MR images, respectively. Compared with multiphasic multidetector CT, gadoxetate disodium-enhanced MR imaging combining dynamic and hepatobiliary phase images results in significantly improved sensitivity and diagnostic accuracy for detection of hypervascular HCC. (orig.)

  20. SPECT/CT and pulmonary embolism

    Energy Technology Data Exchange (ETDEWEB)

    Mortensen, Jann [Copenhagen University Hospital, Department of Clinical Physiology, Nuclear Medicine and PET, Rigshospitalet, Copenhagen (Denmark); The Faroese National Hospital, Department of Medicine, Torshavn (Faroe Islands); Gutte, Henrik [Copenhagen University Hospital, Department of Clinical Physiology, Nuclear Medicine and PET, Rigshospitalet, Copenhagen (Denmark); Herlev Hospital, Copenhagen University Hospital, Department of Radiology, Copenhagen (Denmark); University of Copenhagen, Cluster for Molecular Imaging, Faculty of Health Sciences, Copenhagen (Denmark)

    2014-05-15

    Acute pulmonary embolism (PE) is diagnosed either by ventilation/perfusion (V/P) scintigraphy or pulmonary CT angiography (CTPA). In recent years both techniques have improved. Many nuclear medicine centres have adopted the single photon emission CT (SPECT) technique as opposed to the planar technique for diagnosing PE. SPECT has been shown to have fewer indeterminate results and a higher diagnostic value. The latest improvement is the combination of a low-dose CT scan with a V/P SPECT scan in a hybrid tomograph. In a study comparing CTPA, planar scintigraphy and SPECT alone, SPECT/CT had the best diagnostic accuracy for PE. In addition, recent developments in the CTPA technique have made it possible to image the pulmonary arteries of the lungs in one breath-hold. This development is based on the change from a single-detector to multidetector CT technology with an increase in volume coverage per rotation and faster rotation. Furthermore, the dual energy CT technique is a promising modality that can provide functional imaging in combination with anatomical information. Newer high-end CT scanners and SPECT systems are able to visualize smaller subsegmental emboli. However, consensus is lacking regarding the clinical impact and treatment. In the present review, SPECT and SPECT in combination with low-dose CT, CTPA and dual energy CT are discussed in the context of diagnosing PE. (orig.)

  1. A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT). Part II: On 3D model accuracy

    International Nuclear Information System (INIS)

    Liang Xin; Lambrichts, Ivo; Sun Yi; Denis, Kathleen; Hassan, Bassam; Li Limin; Pauwels, Ruben; Jacobs, Reinhilde

    2010-01-01

    Aim: The study aim was to compare the geometric accuracy of three-dimensional (3D) surface model reconstructions between five Cone Beam Computed Tomography (CBCT) scanners and one Multi-Slice CT (MSCT) system. Materials and methods: A dry human mandible was scanned with five CBCT systems (NewTom 3G, Accuitomo 3D, i-CAT, Galileos, Scanora 3D) and one MSCT scanner (Somatom Sensation 16). A 3D surface bone model was created from the six systems. The reference (gold standard) 3D model was obtained with a high resolution laser surface scanner. The 3D models from the five systems were compared with the gold standard using a point-based rigid registration algorithm. Results: The mean deviation from the gold standard for MSCT was 0.137 mm and for CBCT were 0.282, 0.225, 0.165, 0.386 and 0.206 mm for the i-CAT, Accuitomo, NewTom, Scanora and Galileos, respectively. Conclusion: The results show that the accuracy of CBCT 3D surface model reconstructions is somewhat lower but acceptable comparing to MSCT from the gold standard.

  2. Diagnosis of small posterior fossa stroke on brain CT: effect of iterative reconstruction designed for brain CT on detection performance

    International Nuclear Information System (INIS)

    Inoue, Taihei; Yoshida, Morikatsu; Yokoyama, Koichi; Nakaura, Takeshi; Hirata, Kenichiro; Kidoh, Masafumi; Oda, Seitaro; Utsunomiya, Daisuke; Yamashita, Yasuyuki; Harada, Kazunori

    2017-01-01

    In this study, we aimed to determine whether iterative model reconstruction designed for brain CT (IMR-neuro) would improve the accuracy of posterior fossa stroke diagnosis on brain CT. We enrolled 37 patients with ischaemic stroke in the posterior fossa and 37 patients without stroke (controls). Using axial images reconstructed using filtered back-projection (FBP) and IMR-neuro, we compared the CT numbers in infarcted areas, image noise in the pons, and contrast-to-noise ratios (CNRs) of infarcted and non-infarcted areas on scans subjected to IMR-neuro and FBP. To analyse the performance of hypo-attenuation detection, we used receiver-operating characteristic (ROC) curve techniques. The image noise was significantly lower (2.2 ± 0.5 vs. 5.1 ± 0.9 Hounsfield units, p < 0.01) and the difference in CNR between the infarcted and non-infarcted areas was significantly higher with IMR-neuro than with FBP (2.2 ± 1.7 vs. 4.0 ± 3.6, p < 0.01). Furthermore, the average area under the ROC curve was significantly higher with IMR-neuro (0.90 vs. 0.86 for FBP, p = 0.04). IMR-neuro yielded better image quality and improved hypo-attenuation detection in patients with ischaemic stroke. (orig.)

  3. Diagnosis of small posterior fossa stroke on brain CT: effect of iterative reconstruction designed for brain CT on detection performance

    Energy Technology Data Exchange (ETDEWEB)

    Inoue, Taihei; Yoshida, Morikatsu; Yokoyama, Koichi [Amakusa Medical Center, Department of Radiology, Amakusa, Kumamoto (Japan); Nakaura, Takeshi; Hirata, Kenichiro; Kidoh, Masafumi; Oda, Seitaro; Utsunomiya, Daisuke; Yamashita, Yasuyuki [Kumamoto University, Department of Diagnostic Radiology, Graduate School of Life Sciences, Kumamoto (Japan); Harada, Kazunori [Amakusa Medical Center, Department of Surgery, Kumamoto (Japan)

    2017-09-15

    In this study, we aimed to determine whether iterative model reconstruction designed for brain CT (IMR-neuro) would improve the accuracy of posterior fossa stroke diagnosis on brain CT. We enrolled 37 patients with ischaemic stroke in the posterior fossa and 37 patients without stroke (controls). Using axial images reconstructed using filtered back-projection (FBP) and IMR-neuro, we compared the CT numbers in infarcted areas, image noise in the pons, and contrast-to-noise ratios (CNRs) of infarcted and non-infarcted areas on scans subjected to IMR-neuro and FBP. To analyse the performance of hypo-attenuation detection, we used receiver-operating characteristic (ROC) curve techniques. The image noise was significantly lower (2.2 ± 0.5 vs. 5.1 ± 0.9 Hounsfield units, p < 0.01) and the difference in CNR between the infarcted and non-infarcted areas was significantly higher with IMR-neuro than with FBP (2.2 ± 1.7 vs. 4.0 ± 3.6, p < 0.01). Furthermore, the average area under the ROC curve was significantly higher with IMR-neuro (0.90 vs. 0.86 for FBP, p = 0.04). IMR-neuro yielded better image quality and improved hypo-attenuation detection in patients with ischaemic stroke. (orig.)

  4. Dynamic CT in dissecting aortic aneurysm

    International Nuclear Information System (INIS)

    Fujita, Nobuyuki; Matsuoka, Yosuke; Mizuo, Hideyo; Shirato, Hiroki; Irie, Goro

    1985-01-01

    especially about the branching of major vessels Eight cases of aortic dissection were analysed to evaluate detectability of the branching of major vessels from a ture or false lumen, by dynamic CT. Although recognition of the branching is easy when the direct continuity to a dissected lumen is demonstrated, but without visualization of direct continuity or with no information about dissected lumens, it is sometimes difficult or leads us to misdiagnosis. The diagnostic accuracy is 100% when direct continuity to a dissected lumen is demonstrated, or dynamic CT revealed branching pattern distinctly. Dynamic CT is useful for a diagnosis of dissecting aneurysm and branching pattern of major vessels. (author)

  5. CT Fluoroscopy-Guided Core Biopsy for Diagnosis of Small ({<=} 20 mm) Pulmonary Nodules

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Hye Larn; Kim, Yoon Kyung; Woo, Ok Hee; Yong, Hwan Seok; Kang, Eun Young [Dept. of Radiology, Korea University Guro Hospital, Korea University College of Medicine, Seoul (Korea, Republic of); Kim, Hyun Koo [Dept. of Thoracic Surgery, Korea University Guro Hospital, Korea University College of Medicine, Seoul (Korea, Republic of); Shin, Bong Kyung [Dept. of Pathology, Korea University Guro Hospital, Korea University College of Medicine, Seoul (Korea, Republic of)

    2011-10-15

    To evaluate the efficacy of CT fluoroscopy-guided core biopsy of small pulmonary nodules. This study included 62 patients (35 men, 27 women; age range, 36-85 years) that had a small ({<=} 20 mm) pulmonary nodule and underwent CT fluoroscopy-guided core biopsy. The overall diagnostic accuracy and complication rate were calculated. The diagnostic accuracy was compared between two groups according to the nodule size ({<=} 10 mm vs. > 10 mm), and nodule density (solid vs. subsolid). Malignant or premalignant lesions were finally diagnosed in 39 patients; 36 true-positive and three false-negative findings (sensitivity, 92%). A benign lesion was finally diagnosed in 23 patients, with no false-positive results (specificity, 100%). The overall diagnostic accuracy was 95%. The sensitivity and diagnostic accuracy were 85% and 91% for nodules {<=} 10 mm, and 96% and 97% for nodules > 10 mm (p > 0.05). The sensitivity and diagnostic accuracy were 93% and 96% in the solid group and 90% and 92% in the subsolid group (p > 0.05). Seventeen (27%) patients had a pneumothorax and two (3%) required a closed thoracostomy. CT fluoroscopy-guided core biopsy of small pulmonary nodules yields high diagnostic accuracy with acceptable complication rates.

  6. Technical Note: Insertion of digital lesions in the projection domain for dual-source, dual-energy CT.

    Science.gov (United States)

    Ferrero, Andrea; Chen, Baiyu; Li, Zhoubo; Yu, Lifeng; McCollough, Cynthia

    2017-05-01

    To compare algorithms performing material decomposition and classification in dual-energy CT, it is desirable to know the ground truth of the lesion to be analyzed in real patient data. In this work, we developed and validated a framework to insert digital lesions of arbitrary chemical composition into patient projection data acquired on a dual-source, dual-energy CT system. A model that takes into account beam-hardening effects was developed to predict the CT number of objects with known chemical composition. The model utilizes information about the x-ray energy spectra, the patient/phantom attenuation, and the x-ray detector energy response. The beam-hardening model was validated on samples of iodine (I) and calcium (Ca) for a second-generation dual-source, dual-energy CT scanner for all tube potentials available and a wide range of patient sizes. The seven most prevalent mineral components of renal stones were modeled and digital stones were created with CT numbers computed for each patient/phantom size and x-ray energy spectra using the developed beam-hardening model. Each digital stone was inserted in the dual-energy projection data of a water phantom scanned on a dual-source scanner and reconstructed with the routine algorithms in use in our practice. The geometry of the forward projection for dual-energy data was validated by comparing CT number accuracy and high-contrast resolution of simulated dual-energy CT data of the ACR phantom with experimentally acquired data. The beam-hardening model and forward projection method accurately predicted the CT number of I and Ca over a wide range of tube potentials and phantom sizes. The images reconstructed after the insertion of digital kidney stones were consistent with the images reconstructed from the scanner, and the CT number ratios for different kidney stone types were consistent with data in the literature. A sample application of the proposed tool was also demonstrated. A framework was developed and validated

  7. Regional nodal staging with 18F-FDG PET–CT in non-small cell lung cancer: Additional diagnostic value of CT attenuation and dual-time-point imaging

    International Nuclear Information System (INIS)

    Li, Meng; Wu, Ning; Liu, Ying; Zheng, Rong; Liang, Ying; Zhang, Wenjie; Zhao, Ping

    2012-01-01

    Background: [Fluorine-18]-fluorodeoxyglucose positron emission tomography/computed tomography ( 18 F-FDG PET–CT) is widely performed in the regional nodal staging of non-small cell lung cancer (NSCLC). However, the uptake of 18 F-FDG by tubercular granulomatous tissues may lead to false-positive diagnosis. This is of special concern in China, where tubercular granulomatous disease is epidemic. Herein, we evaluated the efficacy of an additional CT attenuation and a dual-time-point scan in determining the status of lymph nodes. Methods: Eighty NSCLC patients underwent curative surgical resection after 18 F-FDG PET–CT and separate breath-hold CT examinations. The initial images were analyzed by two methods. In method 1, nodal status was determined by 18 F-FDG uptake only. In Method 2, nodal status was determined by 18 F-FDG uptake associated with CT attenuation. For dual-time-point imaging, the retention index (RI) of benign and malignant nodal groups with positive uptake in the initial scan was examined. Results: A total of 265 nodal groups were documented. On a per-nodal-group basis, the diagnostic sensitivity, specificity, and accuracy of Method 1 were 66.7%, 89.7%, and 85.3%, respectively, whereas those of Method 2 were 64.7%, 96.7%, and 90.6%, respectively. The improvement in diagnostic specificity and accuracy associated with the addition of CT attenuation in Method 2 as compared to Method 1 was statistically significant (p 0.05). Conclusion: 18 F-FDG PET–CT has high diagnostic value for preoperative lymph-node (N) staging of NSCLC patients. We show that 18 F-FDG uptake combined with CT attenuation improves the diagnostic specificity and accuracy of nodal diagnosis in NSCLC. For the lymph nodes with positive uptake in the initial scan, dual-time-point imaging has limited effect in differentiation.

  8. Free-breathing cine CT for the diagnosis of tracheomalacia in young children

    International Nuclear Information System (INIS)

    Goo, Hyun Woo

    2013-01-01

    Tracheomalacia is characterized by excessive expiratory collapse of the trachea. To investigate the accuracy of free-breathing cine CT for diagnosis of tracheomalacia in young children with bronchoscopy as reference standard. In a retrospective study (May 2001-July 2008), a patient group (n = 27) of children with bronchoscopic evidence of tracheomalacia, and a control group (n = 320) underwent free-breathing cine CT. The tracheal shape on free-breathing cine CT was classified as round, lunate, elongated or crescentic. Cross-sectional area change of the trachea and age were compared between the groups and the diagnostic performance of free-breathing cine CT for tracheomalacia was evaluated. The patient group showed significantly greater cross-sectional area change of the trachea (57.2% ± 22.2% vs. 10.6% ± 11.2%, P < 0.001) than the control group. If a cross-sectional area change of the trachea of 31.6% was used as a cut-off value for the diagnosis of tracheomalacia, the sensitivity, specificity and accuracy of cine CT were 96.3% (26/27), 97.2% (311/320) and 97.1% (337/347), respectively. If a crescentic shape during the expiratory phase was used, the sensitivity, specificity and accuracy were 51.9% (14/27), 98.8% (316/320) and 95.1% (330/347), respectively. Free-breathing cine CT has potential to provide the diagnosis of tracheomalacia in young children. (orig.)

  9. Spect-CT and PET: CT in the management of differentiated thyroid cancer

    International Nuclear Information System (INIS)

    Khan, Shoukat H.

    2007-01-01

    Full text: With the advancement and evolution in medical technology notably imaging there has been a sea change in the overall management strategy of most of the cancers of human body. The present day functional, imaging with PET and MRI enables us to pick up a tumour at its cellular stage. Molecular imaging and nanotechnology have further added to this expanding armamentarium of cancer imaging and treatment. Thyroid cancer is one such cancer where the cutting edge biotechnology has dramatically changed the management profile of a disease. Thyroid cancer can safely be classified as one of the cancers which if optimally managed is curable. Hybrid and fusion imaging like SPECT-CT and PET-CT with their superior sensitivity and specificity have greatly improved the accuracy of disease detection and reduced drastically the false positive disease sites. Disease not detected by conventional planar imaging can now be detected and also anatomically localized using hybrid imaging modalities of SPECT-CT and PET-CT. An accurate detection and precise localization improves image interpretation and a treatment optimization in the curable cancer of thyroid

  10. Clinical applications of SPECT-CT

    Energy Technology Data Exchange (ETDEWEB)

    Ahmadzadehfar, Hojjat; Biersack, Hans-Juergen (eds.) [University Hospital Bonn (Germany). Dept. of Nuclear Medicine

    2014-06-01

    Covers the full spectrum of clinical applications of SPECT/CT in diagnosis of benign and malignant diseases. Includes chapters on the use of SPECT/CT for dosimetry and for therapy planning. Completely up to date. Many helpful illustrations. SPECT/CT cameras have considerably improved diagnostic accuracy in recent years. Such cameras allow direct correlation of anatomic and functional information, resulting in better localization and definition of scintigraphic findings. In addition to this anatomic referencing, CT coregistration provides superior quantification of radiotracer uptake based on the attenuation correction capabilities of CT. Useful applications of SPECT/CT have been identified not only in oncology but also in other specialties such as orthopedics and cardiology. This book covers the full spectrum of clinical applications of SPECT/CT in diagnosis and therapy planning of benign and malignant diseases. Opening chapters discuss the technology and physics of SPECT/CT and its use for dosimetry. The role of SPECT/CT in the imaging of a range of pathologic conditions is then addressed in detail. Applications covered include, among others, imaging of the thyroid, bone, and lungs, imaging of neuroendocrine tumors, cardiac scintigraphy, and sentinel node scintigraphy. Individual chapters are also devoted to therapy planning in selective internal radiation therapy of liver tumors and bremsstrahlung SPECT/CT. Readers will find this book to be an essential and up-to-date source of information on this invaluable hybrid imaging technique.

  11. The relationship between post-traumatic ossicular injuries and conductive hearing loss: A 3D-CT study.

    Science.gov (United States)

    Maillot, Olivier; Attyé, Arnaud; Boutet, Claire; Boubagra, Kamel; Perolat, Romain; Zanolla, Marion; Grand, Sylvie; Schmerber, Sébastien; Krainik, Alexandre

    2017-09-01

    After a trauma, the conductive ossicular chain may be disrupted by ossicular luxation or fracture. Recent developments in 3D-CT allow a better understanding of ossicular injuries. In this retrospective study, we compared patients with post-traumatic conductive hearing loss (CHL) with those referred without CHL to evaluate the relationship between ossicular injuries and CHL. We also assessed the added value of 3D reconstructions on 2D-CT scan to detect ossicular lesions in patients surgically managed. The CT scans were performed using a 40-section spiral CT scanner in 49 patients with post-traumatic CHL (n=29) and without CHL (n=20). Three radiologists performed independent blind evaluations of 2D-CT and 3D reconstructions to detect ossicular chain injury. We used the t-test to explore differences regarding the number of subjects with ossicular injury in the two groups. We also estimated the diagnostic accuracy and the inter-rater agreement of the 3D-CT reconstructions associated to 2D-CT scan. We identified ossicular abnormality in 14 patients out of 29 and in one patient out of 20 in the CHL and non-CHL groups respectively. There was a significant difference regarding the number of subjects with ossicular lesions between the two groups (P≤0.01). The diagnostic sensitivity of 3D-CT reconstructions associated with 2D-CT ranged from 66% to 100% and the inter-reader agreement ranged from 0.85 to 1, depending of the type of lesion. The relationship between ossicular lesion and the presence of CHL tightly correlated. 3D-CT reconstructions of the temporal bone are useful to assess patients in a post-traumatic context. Copyright © 2017 Elsevier Masson SAS. All rights reserved.

  12. Preoperative CT evaluation of adenocarcinoma of the gastroesophageal junction

    International Nuclear Information System (INIS)

    Bennett, J.D.; Lefcoe, M.S.; Finley, R.; Yoshi, C.; Inculet, R.

    1988-01-01

    A retrospective review was undertaken of 53 preoperative computed tomographic (CT) scans obtained between March 1983 and April 1988 from patients undergoing surgery for adenocarcinoma of the gastroesophageal junction, and results were correlated with the surgical-pathologic findings. CT was unreliable in predicting aortic, pericardial, or pancreatic invasion (sensitivity, 0/8; specificity, 41/45). Of 45 pathologically positive nodal groups, the largest node measured on CT scans was 10 mm or less in 36 cases. The accuracy of preoperative CT in staging adenocarcinoma of the gastroesophageal junction is limited by its low sensitivity in detecting local invasion. Nodal size as measured with CT is not a reliable indicator of metastatic disease

  13. Low Accuracy of Computed Tomography and Positron Emission Tomography to Detect Lung and Lymph Node Metastases of Colorectal Cancer.

    Science.gov (United States)

    Guerrera, Francesco; Renaud, Stéphane; Schaeffer, Mickaël; Nigra, Victor; Solidoro, Paolo; Santelmo, Nicola; Filosso, Pier Luigi; Falcoz, Pierre-Emmanuel; Ruffini, Enrico; Oliaro, Alberto; Massard, Gilbert

    2017-10-01

    Minimally invasive surgery, stereotactic radiotherapy, and radiofrequency ablation are commonly proposed in the case of pulmonary colorectal-metastasis as alternatives to conventional open surgery. Preoperative imaging assessment by computed tomography (CT) scan and fluorodeoxyglucose positron emission tomography (FDG-PET) are critical to guide oncologic radical treatment. Our aim was to investigate the accuracy of CT and FDG-PET for the evaluation of the number of pulmonary colorectal metastases and thoracic lymph nodal involvement (LNI). Patients who underwent lung surgical resection for pulmonary colorectal metastases from 2004 to 2014 were analyzed. Concordance between histology, CT scan, and FDG-PET findings were assessed. Data of 521 patients were analyzed. Of those, FDG-PET was performed in 435 (83.5%). A moderate agreement between both CT scan (kappa index: 0.42) and FDG-PET (kappa index: 0.42) findings and the histologically proven number of metastases was observed. The number of histologically proven metastases was correctly discriminated in 61.7% of cases with CT scan and in 61.8% of cases with FDG-PET. Multiple metastases were discovered in 20.9% of clinical single metastasis cases with CT scan, and in 24.4% of those cases with FDG-PET. One hundred fifty patients (29.1%) presented with pathologic LNI. A poor agreement was observed between LNI and CT scan findings (kappa index: 0.02), and a weak agreement was observed concerning LNI and FDG-PET findings (kappa index: 0.39). Computed tomography and FDG-PET have limitations if the objective is to detect all malignant nodules and to discriminate the LNI in cases of pulmonary metastases of colorectal cancer. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  14. Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors

    International Nuclear Information System (INIS)

    Sweeney, Reinhart A; Seubert, Benedikt; Stark, Silke; Homann, Vanessa; Müller, Gerd; Flentje, Michael; Guckenberger, Matthias

    2012-01-01

    To analyze the accuracy and inter-observer variability of image-guidance (IG) using 3D or 4D cone-beam CT (CBCT) technology in stereotactic body radiotherapy (SBRT) for lung tumors. Twenty-one consecutive patients treated with image-guided SBRT for primary and secondary lung tumors were basis for this study. A respiration correlated 4D-CT and planning contours served as reference for all IG techniques. Three IG techniques were performed independently by three radiation oncologists (ROs) and three radiotherapy technicians (RTTs). Image-guidance using respiration correlated 4D-CBCT (IG-4D) with automatic registration of the planning 4D-CT and the verification 4D-CBCT was considered gold-standard. Results were compared with two IG techniques using 3D-CBCT: 1) manual registration of the planning internal target volume (ITV) contour and the motion blurred tumor in the 3D-CBCT (IG-ITV); 2) automatic registration of the planning reference CT image and the verification 3D-CBCT (IG-3D). Image quality of 3D-CBCT and 4D-CBCT images was scored on a scale of 1–3, with 1 being best and 3 being worst quality for visual verification of the IGRT results. Image quality was scored significantly worse for 3D-CBCT compared to 4D-CBCT: the worst score of 3 was given in 19 % and 7.1 % observations, respectively. Significant differences in target localization were observed between 4D-CBCT and 3D-CBCT based IG: compared to the reference of IG-4D, tumor positions differed by 1.9 mm ± 0.9 mm (3D vector) on average using IG-ITV and by 3.6 mm ± 3.2 mm using IG-3D; results of IG-ITV were significantly closer to the reference IG-4D compared to IG-3D. Differences between the 4D-CBCT and 3D-CBCT techniques increased significantly with larger motion amplitude of the tumor; analogously, differences increased with worse 3D-CBCT image quality scores. Inter-observer variability was largest in SI direction and was significantly larger in IG using 3D-CBCT compared to 4D-CBCT: 0.6 mm versus 1.5 mm

  15. Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors

    Directory of Open Access Journals (Sweden)

    Sweeney Reinhart A

    2012-06-01

    Full Text Available Abstract Background To analyze the accuracy and inter-observer variability of image-guidance (IG using 3D or 4D cone-beam CT (CBCT technology in stereotactic body radiotherapy (SBRT for lung tumors. Materials and methods Twenty-one consecutive patients treated with image-guided SBRT for primary and secondary lung tumors were basis for this study. A respiration correlated 4D-CT and planning contours served as reference for all IG techniques. Three IG techniques were performed independently by three radiation oncologists (ROs and three radiotherapy technicians (RTTs. Image-guidance using respiration correlated 4D-CBCT (IG-4D with automatic registration of the planning 4D-CT and the verification 4D-CBCT was considered gold-standard. Results were compared with two IG techniques using 3D-CBCT: 1 manual registration of the planning internal target volume (ITV contour and the motion blurred tumor in the 3D-CBCT (IG-ITV; 2 automatic registration of the planning reference CT image and the verification 3D-CBCT (IG-3D. Image quality of 3D-CBCT and 4D-CBCT images was scored on a scale of 1–3, with 1 being best and 3 being worst quality for visual verification of the IGRT results. Results Image quality was scored significantly worse for 3D-CBCT compared to 4D-CBCT: the worst score of 3 was given in 19 % and 7.1 % observations, respectively. Significant differences in target localization were observed between 4D-CBCT and 3D-CBCT based IG: compared to the reference of IG-4D, tumor positions differed by 1.9 mm ± 0.9 mm (3D vector on average using IG-ITV and by 3.6 mm ± 3.2 mm using IG-3D; results of IG-ITV were significantly closer to the reference IG-4D compared to IG-3D. Differences between the 4D-CBCT and 3D-CBCT techniques increased significantly with larger motion amplitude of the tumor; analogously, differences increased with worse 3D-CBCT image quality scores. Inter-observer variability was largest in SI direction and was

  16. Accuracy of diagnostic imaging modalities for peripheral post-traumatic osteomyelitis - a systematic review of the recent literature

    Energy Technology Data Exchange (ETDEWEB)

    Govaert, Geertje A. [University of Groningen, University Medical Center Groningen, Department of Surgery, Subdivision of Trauma Surgery, Groningen (Netherlands); University of Utrecht, University Medical Center Utrecht, Department of Trauma Surgery, Utrecht (Netherlands); IJpma, Frank F.; Reininga, Inge H. [University of Groningen, University Medical Center Groningen, Department of Surgery, Subdivision of Trauma Surgery, Groningen (Netherlands); McNally, Martin [Oxford University Hospitals NHS Foundation Trust, The Bone Infection Unit, Nuffield Orthopaedic Centre, Oxford (United Kingdom); McNally, Eugene [Oxford Musculoskeletal Radiology, Oxford (United Kingdom); Glaudemans, Andor W. [University of Groningen, University Medical Center Groningen, Department of Nuclear Medicine and Molecular Imaging, Groningen (Netherlands)

    2017-08-15

    Post-traumatic osteomyelitis (PTO) is difficult to diagnose and there is no consensus on the best imaging strategy. The aim of this study is to present a systematic review of the recent literature on diagnostic imaging of PTO. A literature search of the EMBASE and PubMed databases of the last 16 years (2000-2016) was performed. Studies that evaluated the accuracy of magnetic resonance imaging (MRI), three-phase bone scintigraphy (TPBS), white blood cell (WBC) or antigranulocyte antibody (AGA) scintigraphy, fluorodeoxyglucose positron emission tomography (FDG-PET) and plain computed tomography (CT) in diagnosing PTO were considered for inclusion. The review was conducted using the PRISMA statement and QUADAS-2 criteria. The literature search identified 3358 original records, of which 10 articles could be included in this review. Four of these studies had a comparative design which made it possible to report the results of, in total, 17 patient series. WBC (or AGA) scintigraphy and FDG-PET exhibit good accuracy for diagnosing PTO (sensitivity ranged from 50-100%, specificity ranged from 40-97% versus 83-100% and 51%-100%, respectively). The accuracy of both modalities improved when a hybrid imaging technique (SPECT/CT and FDG-PET/CT) was performed. For FDG-PET/CT, sensitivity ranged between 86 and 94% and specificity between 76 and 100%. For WBC scintigraphy + SPECT/CT, this is 100% and 89-97%, respectively. Based on the best available evidence of the last 16 years, both WBC (or AGA) scintigraphy combined with SPECT/CT or FDG-PET combined with CT have the best diagnostic accuracy for diagnosing peripheral PTO. (orig.)

  17. Fully automated deformable registration of breast DCE-MRI and PET/CT

    Science.gov (United States)

    Dmitriev, I. D.; Loo, C. E.; Vogel, W. V.; Pengel, K. E.; Gilhuijs, K. G. A.

    2013-02-01

    Accurate characterization of breast tumors is important for the appropriate selection of therapy and monitoring of the response. For this purpose breast imaging and tissue biopsy are important aspects. In this study, a fully automated method for deformable registration of DCE-MRI and PET/CT of the breast is presented. The registration is performed using the CT component of the PET/CT and the pre-contrast T1-weighted non-fat suppressed MRI. Comparable patient setup protocols were used during the MRI and PET examinations in order to avoid having to make assumptions of biomedical properties of the breast during and after the application of chemotherapy. The registration uses a multi-resolution approach to speed up the process and to minimize the probability of converging to local minima. The validation was performed on 140 breasts (70 patients). From a total number of registration cases, 94.2% of the breasts were aligned within 4.0 mm accuracy (1 PET voxel). Fused information may be beneficial to obtain representative biopsy samples, which in turn will benefit the treatment of the patient.

  18. Objective image characterization of a spectral CT scanner with dual-layer detector

    Science.gov (United States)

    Ozguner, Orhan; Dhanantwari, Amar; Halliburton, Sandra; Wen, Gezheng; Utrup, Steven; Jordan, David

    2018-01-01

    This work evaluated the performance of a detector-based spectral CT system by obtaining objective reference data, evaluating attenuation response of iodine and accuracy of iodine quantification, and comparing conventional CT and virtual monoenergetic images in three common phantoms. Scanning was performed using the hospital’s clinical adult body protocol. Modulation transfer function (MTF) was calculated for a tungsten wire and visual line pair targets were evaluated. Image noise power spectrum (NPS) and pixel standard deviation were calculated. MTF for monoenergetic images agreed with conventional images within 0.05 lp cm-1. NPS curves indicated that noise texture of 70 keV monoenergetic images is similar to conventional images. Standard deviation measurements showed monoenergetic images have lower noise except at 40 keV. Mean CT number and CNR agreed with conventional images at 75 keV. Measured iodine concentration agreed with true concentration within 6% for inserts at the center of the phantom. Performance of monoenergetic images at detector based spectral CT is the same as, or better than, that of conventional images. Spectral acquisition and reconstruction with a detector based platform represents the physical behaviour of iodine as expected and accurately quantifies the material concentration.

  19. CT image construction of a totally deflated lung using deformable model extrapolation

    International Nuclear Information System (INIS)

    Sadeghi Naini, Ali; Pierce, Greg; Lee, Ting-Yim

    2011-01-01

    intensity mean absolute difference between these two images was calculated to be at 1%. Tumor center as well as a number of anatomical fiducial markers were traced in different corresponding slices of the two images. The average misalignment obtained for the constructed CT image was (0.64, 0.39, 0.11) mm, which indicates a very desirable accuracy for lung brachytherapy applications. Conclusions: The image construction accuracy obtained in this research is suitable for intraoperative tasks; e.g., tumor localization and fusing with real time navigation data in lung brachytherapy. These applications involve image registration with intraoperative U.S. images in order to enhance their poor quality. The proposed technique is also useful for preoperative tasks such as planning of lung brachytherapy treatment.

  20. C-arm cone-beam CT virtual navigation-guided percutaneous mediastinal mass biopsy: Diagnostic accuracy and complications

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Hyungjin [Seoul National University College of Medicine, Department of Radiology, Seoul (Korea, Republic of); Seoul National University Medical Research Center, Institute of Radiation Medicine, Seoul (Korea, Republic of); Aerospace Medical Group, Air Force Education and Training Command, Jinju (Korea, Republic of); Park, Chang Min; Goo, Jin Mo [Seoul National University College of Medicine, Department of Radiology, Seoul (Korea, Republic of); Seoul National University Medical Research Center, Institute of Radiation Medicine, Seoul (Korea, Republic of); Seoul National University, Cancer Research Institute, Seoul (Korea, Republic of); Lee, Sang Min [Seoul National University College of Medicine, Department of Radiology, Seoul (Korea, Republic of); Seoul National University Medical Research Center, Institute of Radiation Medicine, Seoul (Korea, Republic of)

    2015-12-15

    To assess the usefulness of C-arm cone-beam computed tomography (CBCT) virtual navigation-guided percutaneous mediastinal mass biopsy in terms of diagnostic accuracy and complication rates. Seventy-eight CBCT virtual navigation-guided percutaneous mediastinal mass biopsies were performed in 75 patients (M:F, 38:37; mean age, 48.55 ± 18.76 years). The procedural details, diagnostic sensitivity, specificity, accuracy and complication rate were investigated. Mean lesion size was 6.80 ± 3.08 cm, skin-to-target distance was 3.67 ± 1.80 cm, core needle biopsy rate was 96.2 % (75/78), needle indwelling time was 9.29 ± 4.34 min, total procedure time was 13.26 ± 5.29 min, number of biopsy specimens obtained was 3.13 ± 1.02, number of CBCTs performed was 3.03 ± 0.68, rate of lesion border discrimination from abutting mediastinal structures on CBCT was 26.9 % (21/78), technical success rate was 100 % (78/78), estimated effective dose was 5.33 ± 4.99 mSv, and the dose area product was 12,723.68 ± 10,665.74 mGy.cm{sup 2}. Among the 78 biopsies, 69 were malignant, 7 were benign and 2 were indeterminate. Diagnostic sensitivity, specificity and accuracy for the diagnosis of malignancies were 97.1 % (67/69), 100 % (7/7) and 97.4 % (74/76), respectively, with a complication rate of 3.85 % (3/78), all of which were small pneumothoraces. CBCT virtual navigation-guided biopsy is a highly accurate and safe procedure for the evaluation of mediastinal lesions. (orig.)

  1. Flat-detector computed tomography (FD-CT)

    International Nuclear Information System (INIS)

    Kalender, Willi A.; Kyriakou, Yiannis

    2007-01-01

    Flat-panel detectors or, synonymously, flat detectors (FDs) have been developed for use in radiography and fluoroscopy with the defined goal to replace standard X-ray film, film-screen combinations and image intensifiers by an advanced sensor system. FD technology in comparison to X-ray film and image intensifiers offers higher dynamic range, dose reduction, fast digital readout and the possibility for dynamic acquisitions of image series, yet keeping to a compact design. It appeared logical to employ FD designs also for computed tomography (CT) imaging. Respective efforts date back a few years only, but FD-CT has meanwhile become widely accepted for interventional and intra-operative imaging using C-arm systems. FD-CT provides a very efficient way of combining two-dimensional (2D) radiographic or fluoroscopic and 3D CT imaging. In addition, FD technology made its way into a number of dedicated CT scanner developments, such as scanners for the maxillo-facial region or for micro-CT applications. This review focuses on technical and performance issues of FD technology and its full range of applications for CT imaging. A comparison with standard clinical CT is of primary interest. It reveals that FD-CT provides higher spatial resolution, but encompasses a number of disadvantages, such as lower dose efficiency, smaller field of view and lower temporal resolution. FD-CT is not aimed at challenging standard clinical CT as regards to the typical diagnostic examinations; but it has already proven unique for a number of dedicated CT applications, offering distinct practical advantages, above all the availability of immediate CT imaging in the interventional suite or the operating room. (orig.)

  2. Development of a coupled tendon-driven 3D multi-joint manipulator. Investigation of tension transfer efficiency, optimal reel arrangement and tip positioning accuracy

    International Nuclear Information System (INIS)

    Horigome, Atsushi; Yamada, Hiroya; Hirose, Shigeo; Sen, Shin; Endo, Gen

    2017-01-01

    Long-reach robotic manipulators are expected to be used in the space where humans cannot work such as nuclear power plant disaster areas. We suggested a coupled tendon-driven articulated manipulator '3D CT-Arm' and developed a preliminary prototype 'Mini 3D CT-Arm' whose arm had 2.4 m length and 0.3 m width. In order to consider developing '3D CT-Arm' deeply, we discussed tension transfer efficiency of a tendon through pulleys, the arrangement of the maximum number of reels in a limited space and the tip positioning accuracy. Through many transfer efficiency experiments, we conclude that tension transfer efficiency of '3D CT-Arm' can reach over 88% in the worst case. We investigated non-interfering reels' arrangement in the base by full search in cases of up to 10 reels. In all simulations, V-shaped or W-shaped arrangement can support the most reels in a limited space. Therefore, we conclude this is the most optimal reels' arrangement. Finally, we carried out the positioning accuracy experiment with 'Mini 3D CT-Arm' via motion capture system. Although the tip position had a 2 to 41 mm error between the desired value and the measured value by potentiometer, a 29 to 95 mm error between the desired value and the measured value was measured by motion capture system. (author)

  3. Rectosigmoid endometriosis: comparison between CT water enema and video laparoscopy.

    Science.gov (United States)

    Stabile Ianora, A A; Moschetta, M; Lorusso, F; Lattarulo, S; Telegrafo, M; Rella, L; Scardapane, A

    2013-09-01

    To evaluate the accuracy of water enema computed tomography (CT) for predicting the location of endometriosis in patients with contraindications to magnetic resonance imaging (MRI), focusing on rectosigmoid lesions and having laparoscopic and histological data as the reference standard. Thirty-three women (mean age 33.4 ± 3.1 years) suspected of having deep pelvic endometriosis underwent 64-row CT and video laparoscopy within 4 weeks. Two radiologists blinded to the clinical data evaluated the CT images obtained after colonic retrograde distension using water as the contrast medium, and a comparison with laparoscopic and histological findings was performed. CT sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and diagnostic accuracy were calculated. The radiation dose to patients was estimated. Cohen's weighted kappa (κ) test was used to evaluate the interobserver agreement. In 23 out of 33 patients (69%) intestinal implants were found at surgery and pathological examinations. CT confirmed the diagnosis of rectosigmoid endometriosis in 20 out of 23 implants. Three nodules located on the proximal sigmoid colon (two serosal lesions and one infiltrating the muscularis layer) with a diameter of less than 1 cm were not diagnosed. CT sensitivity, specificity, PPV, NPV, and accuracy values were 87, 100, 100, 77, and 91%, respectively. The mean effective dose estimate was 6.30 ± 1.7 mSv. Almost perfect agreement between the two readers was found (k = 0.84). Water enema CT can play a role in the diagnosis of bowel endometriosis and represents another accurate potential tool for video laparoscopic approaches, especially in patients for whom MRI is contraindicated. Copyright © 2013 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.

  4. Clinical evaluation of high-resolution CT, 1. CT diagnosis of liver tumors and its limit

    Energy Technology Data Exchange (ETDEWEB)

    Araki, T [Tokyo Univ. (Japan). Faculty of Medicine

    1980-03-01

    To estimate diagnostic accuracy of CT in liver tumors, CT diagnosis in 120 patients with primary hepatocellular carcinoma was discussed. As a result, primary hepatocellular carcinoma less than 2 cm in diameter could not be visualized by CT. Even tumors between 4 and 8 cm in diameter showed false negative caused by isodense tumors on images of 4 patients. To improve the detectability of liver tumors by CT, the higher resolution of low contrast regions on images are required. As a method to improve qualitative diagnosis of liver tumors, rapid intravenous injection of contrast medium was performed on 42 patients with liver tumors, As a result, images reflecting vascularity of tumors were obtained, and the differential diagnosis was possible to some extent by observing the movement of the contrast. Especially, cavernous hemangioma could be distinguished from hepatocellular carcinoma, because cavernous hemangioma showed specific images and could be diagnosed accurately.

  5. Evaluation of living renal donors: accuracy of three-dimensional 16-section CT

    Energy Technology Data Exchange (ETDEWEB)

    Rastogi, N; Sahani, D.V.; Blake, M.A.; Ko, D.C.; Mueller, P.R. [Massachusetts General Hospital, Boston, MA (United States). Dept. of Radiology

    2006-07-15

    Purpose: To retrospectively assess the sensitivity and specificity of three-dimensional (3D) 16-section computed tomography (CT) in the evaluation of vessels, pelvicalyceal system, and ureters in living renal donors, with surgical findings as the reference standard. Materials and methods: This was a HIPAA-compliant study. Institutional review board approval was obtained for the review of subjects' medical records and data analysis, with waiver of informed consent. Forty-six renal donors (18 men, 28 women; mean age, 42 years) were examined with 16-section CT. Two blinded reviewers independently studied renal vascular and urographic anatomy of each donor CT scans by fist using 3D images alone, then transverse images alone, and finally transverse and 3D data set. Image quality, degree of diagnostic confidence, and time used for review were recorded. Sensitivity and specificity were calculated. Results: For 3D images, transverse images, and transverse in conjunction with 3D data sets, the respective sensitivity and specificity of CT in evaluation of accessory arteries by reviewer 1 were 100% and 100%, 89% and 100%, and 100% and 100%, and those by reviewer 2 were 89% and 97%, 89% and 100%, and 89% and 100%; the respective sensitivity and specificity in evaluation of venous anomalies by reviewer 1 were 100% and 98%, 100% and 98%, and 100% and 98%, and those by reviewer 2 were 100% and 98%, 100% and 95%, and 100% and 98%. For focused comprehensive assessment of renal donors with 3D scans alone, a reviewer on average (average of reviewers 1 and 2) used 2.4 minutes per scan, demonstrated full confidence in 93%, and rated the quality as excellent in 76%. Conclusion: For focused assessment of renal vascular and urographic anatomy, review of 3D data set alone provides high sensitivity and specificity with regard to findings seen at surgery. (author)

  6. Evaluation of living renal donors: accuracy of three-dimensional 16-section CT

    International Nuclear Information System (INIS)

    Rastogi, N; Sahani, D.V.; Blake, M.A.; Ko, D.C.; Mueller, P.R.

    2006-01-01

    Purpose: To retrospectively assess the sensitivity and specificity of three-dimensional (3D) 16-section computed tomography (CT) in the evaluation of vessels, pelvicalyceal system, and ureters in living renal donors, with surgical findings as the reference standard. Materials and methods: This was a HIPAA-compliant study. Institutional review board approval was obtained for the review of subjects' medical records and data analysis, with waiver of informed consent. Forty-six renal donors (18 men, 28 women; mean age, 42 years) were examined with 16-section CT. Two blinded reviewers independently studied renal vascular and urographic anatomy of each donor CT scans by fist using 3D images alone, then transverse images alone, and finally transverse and 3D data set. Image quality, degree of diagnostic confidence, and time used for review were recorded. Sensitivity and specificity were calculated. Results: For 3D images, transverse images, and transverse in conjunction with 3D data sets, the respective sensitivity and specificity of CT in evaluation of accessory arteries by reviewer 1 were 100% and 100%, 89% and 100%, and 100% and 100%, and those by reviewer 2 were 89% and 97%, 89% and 100%, and 89% and 100%; the respective sensitivity and specificity in evaluation of venous anomalies by reviewer 1 were 100% and 98%, 100% and 98%, and 100% and 98%, and those by reviewer 2 were 100% and 98%, 100% and 95%, and 100% and 98%. For focused comprehensive assessment of renal donors with 3D scans alone, a reviewer on average (average of reviewers 1 and 2) used 2.4 minutes per scan, demonstrated full confidence in 93%, and rated the quality as excellent in 76%. Conclusion: For focused assessment of renal vascular and urographic anatomy, review of 3D data set alone provides high sensitivity and specificity with regard to findings seen at surgery. (author)

  7. Non-small-cell lung cancer: detection of mediastinal lymph node metastases by endoscopic ultrasound and CT

    International Nuclear Information System (INIS)

    Potepan, P.; Meroni, E.; Spagnoli, I.; Milella, M.; Danesini, G.M.; Laffranchi, A.; Civelli, E.; Alloisio, M.; Mariani, L.; Spinelli, P.; Guzzon, A.

    1996-01-01

    In this prospective study endoscopic ultrasound (EUS) and computed tomography (CT) were evaluated to compare diagnostic accuracy of the two methods. They were performed for nodal staging in selected patients admitted to our institution for non-small-cell lung cancer (NSCLC). From February 1992 to July 1993, 45 patients were recruited for the study when N3 and N2 nodal involvement were excluded on standard chest X-ray. All the patients completed EUS and CT exams for staging before treatment. The results of sensitivity, specificity and accuracy were obtained in 30 patients who underwent surgical treatment with macroscopically radical resection of T and N, which allowed a complete surgical and histological comparison of CT and EUS findings. On a per-patient basis CT results were: Sensitivity 63.6%, specificity 78.9% and accuracy of 73.3%; on a nodal station basis sensitivity, specificity and accuracy were 70.0%, 85,1% and 81.6%, respectively. The EUS evaluation showed, on a per-patient basis, values of sensitivity 45.5%, specificity 57.9% and overall diagnostic accuracy of 53.3%. On a nodal station basis the results were 50.0%, 86.6% and 78.2%, respectively. The results obtained in the 30 patients when both techniques were taken in association regarding senstivity (90.9%), specificity (73.7%) and accuracy (80.0%) on a per-patient basis suggest that the association of EUS and CT offers the best approach for preoperative staging of NSCLC. (orig.)

  8. TH-CD-206-09: Learning-Based MRI-CT Prostate Registration Using Spare Patch-Deformation Dictionary

    International Nuclear Information System (INIS)

    Yang, X; Jani, A; Rossi, P; Mao, H; Curran, W; Liu, T

    2016-01-01

    Purpose: To enable MRI-guided prostate radiotherapy, MRI-CT deformable registration is required to map the MRI-defined tumor and key organ contours onto the CT images. Due to the intrinsic differences in grey-level intensity characteristics between MRI and CT images, the integration of MRI into CT-based radiotherapy is very challenging. We are developing a learning-based registration approach to address this technical challenge. Methods: We propose to estimate the deformation between MRI and CT images in a patch-wise fashion by using the sparse representation technique. Specifically, we assume that two image patches should follow the same deformation if their patch-wise appearance patterns are similar. We first extract a set of key points in the new CT image. Then, for each key point, we adaptively construct a coupled dictionary from the training MRI-CT images, where each coupled element includes both appearance and deformation of the same image patch. After calculating the sparse coefficients in representing the patch appearance of each key point based on the constructed dictionary, we can predict the deformation for this point by applying the same sparse coefficients to the respective deformations in the dictionary. Results: This registration technique was validated with 10 prostate-cancer patients’ data and its performance was compared with the commonly used free-form-deformation-based registration. Several landmarks in both images were identified to evaluate the accuracy of our approach. Overall, the averaged target registration error of the intensity-based registration and the proposed method was 3.8±0.4 mm and 1.9±0.3 mm, respectively. Conclusion: We have developed a novel prostate MR-CT registration approach based on patch-deformation dictionary, demonstrated its clinical feasibility, and validated its accuracy. This technique will either reduce or compensate for the effect of patient-specific treatment variation measured during the course of

  9. TH-CD-206-09: Learning-Based MRI-CT Prostate Registration Using Spare Patch-Deformation Dictionary

    Energy Technology Data Exchange (ETDEWEB)

    Yang, X; Jani, A; Rossi, P; Mao, H; Curran, W; Liu, T [Emory University, Atlanta, GA (United States)

    2016-06-15

    Purpose: To enable MRI-guided prostate radiotherapy, MRI-CT deformable registration is required to map the MRI-defined tumor and key organ contours onto the CT images. Due to the intrinsic differences in grey-level intensity characteristics between MRI and CT images, the integration of MRI into CT-based radiotherapy is very challenging. We are developing a learning-based registration approach to address this technical challenge. Methods: We propose to estimate the deformation between MRI and CT images in a patch-wise fashion by using the sparse representation technique. Specifically, we assume that two image patches should follow the same deformation if their patch-wise appearance patterns are similar. We first extract a set of key points in the new CT image. Then, for each key point, we adaptively construct a coupled dictionary from the training MRI-CT images, where each coupled element includes both appearance and deformation of the same image patch. After calculating the sparse coefficients in representing the patch appearance of each key point based on the constructed dictionary, we can predict the deformation for this point by applying the same sparse coefficients to the respective deformations in the dictionary. Results: This registration technique was validated with 10 prostate-cancer patients’ data and its performance was compared with the commonly used free-form-deformation-based registration. Several landmarks in both images were identified to evaluate the accuracy of our approach. Overall, the averaged target registration error of the intensity-based registration and the proposed method was 3.8±0.4 mm and 1.9±0.3 mm, respectively. Conclusion: We have developed a novel prostate MR-CT registration approach based on patch-deformation dictionary, demonstrated its clinical feasibility, and validated its accuracy. This technique will either reduce or compensate for the effect of patient-specific treatment variation measured during the course of

  10. Hypothalamic-pituitary dwarfism: Comparison between MR imaging and CT findings

    International Nuclear Information System (INIS)

    Maghnie, M.; Larizza, D.; Severi, F.; Triulzi, F.; Scotti, G.; Beluffi, G.; Cecchini, A.

    1990-01-01

    Magnetic Resonance (MR) imaging was carried out on 33 patients with idiopathic growth hormone deficiency, in 22 of whom CT scan had been carried out previously. Twenty-one patients presented some complications at birth. Both MR and CT were positive in the evaluation of the sella. MR imaging exhibited a higher degree of accuracy than CT in the evaluation of pituitary gland, pituitary stalk and brain anomalies. (orig.)

  11. Patient-specific estimation of detailed cochlear shape from clinical CT images

    DEFF Research Database (Denmark)

    Kjer, H Martin; Fagertun, Jens; Wimmer, Wilhelm

    2018-01-01

    of the detailed patient-specific cochlear shape from CT images. From a collection of temporal bone [Formula: see text]CT images, we build a cochlear statistical deformation model (SDM), which is a description of how a human cochlea deforms to represent the observed anatomical variability. The model is used...... for regularization of a non-rigid image registration procedure between a patient CT scan and a [Formula: see text]CT image, allowing us to estimate the detailed patient-specific cochlear shape. We test the accuracy and precision of the predicted cochlear shape using both [Formula: see text]CT and CT images...

  12. TU-A-12A-12: Improved Airway Measurement Accuracy for Low Dose Quantitative CT (qCT) Using Statistical (ASIR), at Reduced DFOV, and High Resolution Kernels in a Phantom and Swine Model

    Energy Technology Data Exchange (ETDEWEB)

    Yadava, G; Imai, Y; Hsieh, J

    2014-06-15

    Purpose: Quantitative accuracy of Iodine Hounsfield Unit (HU) in conventional single-kVp scanning is susceptible to beam-hardening effect. Dual-energy CT has unique capabilities of quantification using monochromatic CT images, but this scanning mode requires the availability of the state-of-the-art CT scanner and, therefore, is limited in routine clinical practice. Purpose of this work was to develop a beam-hardening-correction (BHC) for single-kVp CT that can linearize Iodine projections at any nominal energy, apply this approach to study Iodine response with respect to keV, and compare with dual-energy based monochromatic images obtained from material-decomposition using 80kVp and 140kVp. Methods: Tissue characterization phantoms (Gammex Inc.), containing solid-Iodine inserts of different concentrations, were scanned using GE multi-slice CT scanner at 80, 100, 120, and 140 kVp. A model-based BHC algorithm was developed where Iodine was estimated using re-projection of image volume and corrected through an iterative process. In the correction, the re-projected Iodine was linearized using a polynomial mapping between monochromatic path-lengths at various nominal energies (40 to 140 keV) and physically modeled polychromatic path-lengths. The beam-hardening-corrected 80kVp and 140kVp images (linearized approximately at effective energy of the beam) were used for dual-energy material-decomposition in Water-Iodine basis-pair followed by generation of monochromatic images. Characterization of Iodine HU and noise in the images obtained from singlekVp with BHC at various nominal keV, and corresponding dual-energy monochromatic images, was carried out. Results: Iodine HU vs. keV response from single-kVp with BHC and dual-energy monochromatic images were found to be very similar, indicating that single-kVp data may be used to create material specific monochromatic equivalent using modelbased projection linearization. Conclusion: This approach may enable quantification of

  13. The diagnostic accuracy of US, CT, MRI and {sup 1}H-MRS for the evaluation of hepatic steatosis compared with liver biopsy: a meta-analysis

    Energy Technology Data Exchange (ETDEWEB)

    Bohte, Anneloes E.; Werven, Jochem R. van; Bipat, Shandra; Stoker, Jaap [University of Amsterdam, Department of Radiology, Academic Medical Center, Amsterdam (Netherlands)

    2011-01-15

    To meta-analyse the diagnostic accuracy of US, CT, MRI and {sup 1}H-MRS for the evaluation of hepatic steatosis. From a comprehensive literature search in MEDLINE, EMBASE, CINAHL and Cochrane (up to November 2009), articles were selected that investigated the diagnostic performance imaging techniques for evaluating hepatic steatosis with histopathology as the reference standard. Cut-off values for the presence of steatosis on liver biopsy were subdivided into four groups: (1) >0, >2 and >5% steatosis; (2) >10, >15 and >20%; (3) >25, >30 and >33%; (4) >50, >60 and >66%. Per group, summary estimates for sensitivity and specificity were calculated. The natural-logarithm of the diagnostic odds ratio (lnDOR) was used as a single indicator of test performance. 46 articles were included. Mean sensitivity estimates for subgroups were 73.3-90.5% (US), 46.1-72.0% (CT), 82.0-97.4% (MRI) and 72.7-88.5% ({sup 1}H-MRS). Mean specificity ranges were 69.6-85.2% (US), 88.1-94.6% (CT), 76.1-95.3% (MRI) and 92.0-95.7% ({sup 1}H-MRS). Overall performance (lnDOR) of MRI and {sup 1}H-MRS was better than that for US and CT for all subgroups, with significant differences in groups 1 and 2. MRI and {sup 1}H-MRS can be considered techniques of choice for accurate evaluation of hepatic steatosis. (orig.)

  14. An assessment of the number of CT slices necessary to plan breast radiotherapy

    International Nuclear Information System (INIS)

    Vincent, D.; Beckham, W.; Delaney, G.

    1999-01-01

    The aim of this study was to evaluate the number of CT slices required to produce satisfactory dose distribution for tangential field irradiation of the chest wall and breast and to assess correlation of this with the volume of breast tissue treated. We conclude that a single-slice plan is unsatisfactory in providing sufficient information about the dose variation across the treatment volume and that ideally a 3D plan with DVHs should be produced. If the required data is unavailable then a minimum of three slices should be used as an approximation. We also propose a software tool for treatment planning systems, which calculates the percentage of the total PTV having dose outside the ICRU 50 radiation dose distribution homogeneity guideline range

  15. Detection of distant metastasis and prognostic prediction of recurrent salivary gland carcinomas using 18 F-FDG PET/CT.

    Science.gov (United States)

    Lee, S H; Roh, J-L; Kim, J S; Lee, J H; Choi, S-H; Nam, S Y; Kim, S Y

    2018-04-24

    To compare the diagnostic accuracy of 18 F-FDG PET/CT and conventional contrast-enhanced CT for the re-staging of recurrent salivary gland carcinoma (SGC). This study included 58 consecutive patients who underwent recurrent SGCs after definitive treatment. The recurrences were evaluated by 18 F-FDG PET/CT and contrast-enhanced CT of the neck and chest. McNemar's test was used to compare the diagnostic accuracy of 18 F-FDG PET/CT with standard neck and chest CT imaging, and a Cox proportional hazards model was used to assess the prognostic value of PET/CT. Of 58 patients with recurrent SGCs, 17 (29%) had a local recurrence, 17 (29%) had a regional recurrence, and 38 (66%) had a distant metastasis, with these classifications showing overlap. The sensitivity and accuracy of 18 F-FDG PET/CT for the detection of distant metastases were significantly higher than those of CT (P 0.1). The 18 F-FDG PET/CT-positive findings at distant sites were predictors of poor progression-free and overall survival outcome (all P PET/CT is a more effective method than CT for detecting distant site recurrences of SGC. This may lead to prognostic prediction for recurrent SGCs. This article is protected by copyright. All rights reserved. This article is protected by copyright. All rights reserved.

  16. Influence of radiation dose and iterative reconstruction algorithms for measurement accuracy and reproducibility of pulmonary nodule volumetry: A phantom study.

    Science.gov (United States)

    Kim, Hyungjin; Park, Chang Min; Song, Yong Sub; Lee, Sang Min; Goo, Jin Mo

    2014-05-01

    To evaluate the influence of radiation dose settings and reconstruction algorithms on the measurement accuracy and reproducibility of semi-automated pulmonary nodule volumetry. CT scans were performed on a chest phantom containing various nodules (10 and 12mm; +100, -630 and -800HU) at 120kVp with tube current-time settings of 10, 20, 50, and 100mAs. Each CT was reconstructed using filtered back projection (FBP), iDose(4) and iterative model reconstruction (IMR). Semi-automated volumetry was performed by two radiologists using commercial volumetry software for nodules at each CT dataset. Noise, contrast-to-noise ratio and signal-to-noise ratio of CT images were also obtained. The absolute percentage measurement errors and differences were then calculated for volume and mass. The influence of radiation dose and reconstruction algorithm on measurement accuracy, reproducibility and objective image quality metrics was analyzed using generalized estimating equations. Measurement accuracy and reproducibility of nodule volume and mass were not significantly associated with CT radiation dose settings or reconstruction algorithms (p>0.05). Objective image quality metrics of CT images were superior in IMR than in FBP or iDose(4) at all radiation dose settings (pvolumetry can be applied to low- or ultralow-dose chest CT with usage of a novel iterative reconstruction algorithm without losing measurement accuracy and reproducibility. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  17. Estimation of mouse organ locations through registration of a statistical mouse atlas with micro-CT images.

    Science.gov (United States)

    Wang, Hongkai; Stout, David B; Chatziioannou, Arion F

    2012-01-01

    Micro-CT is widely used in preclinical studies of small animals. Due to the low soft-tissue contrast in typical studies, segmentation of soft tissue organs from noncontrast enhanced micro-CT images is a challenging problem. Here, we propose an atlas-based approach for estimating the major organs in mouse micro-CT images. A statistical atlas of major trunk organs was constructed based on 45 training subjects. The statistical shape model technique was used to include inter-subject anatomical variations. The shape correlations between different organs were described using a conditional Gaussian model. For registration, first the high-contrast organs in micro-CT images were registered by fitting the statistical shape model, while the low-contrast organs were subsequently estimated from the high-contrast organs using the conditional Gaussian model. The registration accuracy was validated based on 23 noncontrast-enhanced and 45 contrast-enhanced micro-CT images. Three different accuracy metrics (Dice coefficient, organ volume recovery coefficient, and surface distance) were used for evaluation. The Dice coefficients vary from 0.45 ± 0.18 for the spleen to 0.90 ± 0.02 for the lungs, the volume recovery coefficients vary from 0.96 ± 0.10 for the liver to 1.30 ± 0.75 for the spleen, the surface distances vary from 0.18 ± 0.01 mm for the lungs to 0.72 ± 0.42 mm for the spleen. The registration accuracy of the statistical atlas was compared with two publicly available single-subject mouse atlases, i.e., the MOBY phantom and the DIGIMOUSE atlas, and the results proved that the statistical atlas is more accurate than the single atlases. To evaluate the influence of the training subject size, different numbers of training subjects were used for atlas construction and registration. The results showed an improvement of the registration accuracy when more training subjects were used for the atlas construction. The statistical atlas-based registration was also compared with

  18. Determination of the integrated CT number of the whole liver in patients with severe hepatitis. As an indicator of the functional reserve of the liver

    Energy Technology Data Exchange (ETDEWEB)

    Kumahara, Tadashi; Muto, Yasutoshi; Moriwaki, Hisataka; Yoshida, Takashi; Tomita, Eiichi (Gifu Univ. (Japan). Faculty of Medicine)

    1989-06-01

    A study was conducted to estimate the functional reserve of the liver of patients with severe hepatitis by computed tomography (CT), in particular employing the integrated CT number of the whole liver (ICTN). ICTN was calculated by integrating the product of 'area' times 'mean CT number' of the liver in each CT slice for the entire height of the liver. The following results were obtained: (1) In patients with fulminant hepatitis (FH) as well as those with subacute hepatitis (SAH), ICTN was found to be significantly lower as compared to that of patients with acute hepatitis (AH) or non-hepatic diseases. In addition, in FH and SAH patients, ICTN showed a larger degree of decrease when compared with such conventional parameters as either estimated liver volume or mean hepatic CT number. Thus, ICTN seems to more sensitively reflect the changes in functional reserve of the liver. (2) ICTN showed significant positive correlations with prothrombin time and plasma BCAA/AAA ratio, and a significant negative correlation with plasma methionine level. (3) Time course of changes in ICTN correlated well with the clinical features of severe hepatitis. In particular, patients with initial ICTN values above 20 lcenter dotHU/m/sup 2/ of body surface area showed significantly higher survival rate than those with initial ICTN below 20. In conclusion, ICTN well indicates the functional reserve of the liver, and is further suggested to be valuable as a parameter to predict the prognosis of patients with severe hepatitis. (author).

  19. Solitary pulmonary nodules: impact of functional CT on the cost-effectiveness of FDG-PET

    International Nuclear Information System (INIS)

    Miles, K.A.; Keith, C.J.; Wong, D.C.; Griffiths, M.R.

    2002-01-01

    Full text: FDG-PET has been shown to be cost-effective for the evaluation of solitary pulmonary nodules (SPNs) in Australia. This study evaluates the impact on cost-effectiveness produced by incorporating a novel CT technique, functional CT, into diagnostic algorithms for characterisation of SPNs. Four diagnostic strategies were evaluated using decision tree sensitivity analysis. The first strategy comprised patients undergoing conventional CT alone (CT). The second comprised conventional CT followed by functional CT study (FCT), when the SPN was not benign on conventional CT. The third strategy comprised conventional CT, which if positive is followed by FDG-PET (PET) and a fourth strategy where patients with a positive conventional CT undergo functional CT, which if positive also undergo FDG-PET (FCT+PET). Values for disease prevalence and diagnostic accuracy of PET, CT and functional CT were obtained from a literature review, using Australia values where available. Procedure costs were derived from the Medicare Benefits Schedule and DRG Cost Weights for Australian public hospitals. The cost per patient, accuracy and Incremental Cost-Accuracy Ratio (ICAR) were determined for each strategy. Sensitivity analysis evaluated the effect of disease prevalence on cost-effectiveness. Results: At the prevalence of malignancy reported from Australian series (54%), the FCT strategy incurs the least cost ($5560/patient), followed by the FCT+PET ($5910/patient). The FCT+PET strategy is the most cost-effective strategy with an ICAR of $12059/patient, followed by the PET strategy with an ICAR of $12300/patient. At levels of disease prevalence below 54% the above relationship for cost-effectiveness remains the same. For high levels of disease prevalence, CT or FCT are found to be more cost-effective. At typical prevalence of malignancy the cost-effectiveness of PET is enhanced by the addition of functional CT, but at high prevalence functional CT alone is most cost

  20. PET/CT in staging of the high risk prostate cancer

    International Nuclear Information System (INIS)

    Bergero, M.A.; David, C.; Dipatto, F.; Popeneciu, V.; Ríos, L.; Faccio, F.

    2016-01-01

    Objectives: In the last decade multimodal management of the high risk prostate cancer (HRPC) is a therapeutic option in selected patients and the staging of these patients depends on the current diagnostic methods (DM) which have low diagnostic accuracy for detecting metastasis (MTS). The positron emission tomography/computed tomography (PET/CT) would have a greater diagnostic accuracy and it is presented as a better DM for staging prostate cancer (PC). The aim of this article is present 2 patients in whom PET/CT modified the therapeutic decision and conduct a literature review. Materials and methods: 2 patients with HRPC who performed PET/CT and it modified the therapeutic behavior were described and a systematic review of the literature was conducted using PubMed, Embase, SciELO and Cochrane answering the question: has PET/CT a place in HRPC staging? Results: TPET/CT has a sensitivity and specificity between 19% to 100% and 67% to 98,5 %, respectively, in assessing nodal involvement by PC and between 84% to 96% and 92.3% to 100%, respectively, in assessing bone involvement by PC. Besides PET/CT allowed to modify the therapeutic behavior between 20% to 40% of the patients with PC. Conclusions: PET/CT has good specificity and moderate sensitivity for detecting lymph node MTS and good sensitivity and specificity for detecting bone MTS. Besides PET/CT modified the therapeutic behavior in 1/3 of cases and it allowed us to modify the therapeutic behavior in our series. (authors) [es