WorldWideScience

Sample records for calibration standards

  1. Radiological Calibration and Standards Facility

    Data.gov (United States)

    Federal Laboratory Consortium — PNNL maintains a state-of-the-art Radiological Calibration and Standards Laboratory on the Hanford Site at Richland, Washington. Laboratory staff provide expertise...

  2. Performance standard for dose Calibrator

    CERN Document Server

    Darmawati, S

    2002-01-01

    Dose calibrator is an instrument used in hospitals to determine the activity of radionuclide for nuclear medicine purposes. International Electrotechnical Commission (IEC) has published IEC 1303:1994 standard that can be used as guidance to test the performance of the instrument. This paper briefly describes content of the document,as well as explains the assessment that had been carried out to test the instrument accuracy in Indonesia through intercomparison measurement.Its is suggested that hospitals acquire a medical physicist to perform the test for its dose calibrator. The need for performance standard in the form of Indonesia Standard is also touched.

  3. Calibration of surface roughness standards

    DEFF Research Database (Denmark)

    Thalmann, R.; Nicolet, A.; Meli, F.;

    2016-01-01

    organisations. Five surface texture standards of different type were circulated and on each of the standards several roughness parameters according to the standard ISO 4287 had to be determined. 32 out of 395 individual results were not consistent with the reference value. After some corrective actions...... the number of inconsistent results could be reduced to 20, which correspond to about 5% of the total and can statistically be expected. In addition to the material standards, two softgauges were circulated, which allow to test the software of the instruments used in the comparison. The comparison results...... help to support the calibraton and measurement capabilities (CMCs) of the laboratories involved in the CIPM MRA. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final...

  4. Intercomparison of Laboratory Radiance Calibration Standards

    Science.gov (United States)

    Pavri, Betina; Chrien, Tom; Green, Robert; Williams, Orlesa

    2000-01-01

    Several standards for radiometric calibration were measured repeatedly with a spectroradiometer in order to understand how they compared in accuracy and stability. The tested radiance standards included a NIST 1000 W bulb and halon panel, two calibrated and stabilized integrating spheres, and a cavity blackbody. Results indicate good agreement between the blackbody and 1000 W bulb/spectralon panel, If these two radiance sources are assumed correct, then the integrating spheres did not conform. to their manufacturer-reported radiances in several regions of the spectrum. More detailed measurements am underway to investigate the discrepancy.

  5. LLNL X-ray Calibration and Standards Laboratory

    International Nuclear Information System (INIS)

    The LLNL X-ray Calibration and Standards Laboratory is a unique facility for developing and calibrating x-ray sources, detectors, and materials, and for conducting x-ray physics research in support of our weapon and fusion-energy programs

  6. Radioactive standards and calibration methods for contamination monitoring instruments

    Energy Technology Data Exchange (ETDEWEB)

    Yoshida, Makoto [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    1997-06-01

    Contamination monitoring in the facilities for handling unsealed radioactive materials is one of the most important procedures for radiation protection as well as radiation dose monitoring. For implementation of the proper contamination monitoring, radiation measuring instruments should not only be suitable to the purpose of monitoring, but also be well calibrated for the objective qualities of measurement. In the calibration of contamination monitoring instruments, quality reference activities need to be used. They are supplied in different such as extended sources, radioactive solutions or radioactive gases. These reference activities must be traceable to the national standards or equivalent standards. On the other hand, the appropriate calibration methods must be applied for each type of contamination monitoring instruments. In this paper, the concepts of calibration for contamination monitoring instruments, reference sources, determination methods of reference quantities and practical calibration methods of contamination monitoring instruments, including the procedures carried out in Japan Atomic Energy Research Institute and some relevant experimental data. (G.K.)

  7. HPGe Detector Efficiency Calibration Using HEU Standards

    Energy Technology Data Exchange (ETDEWEB)

    Salaymeh, S.R.

    2000-10-12

    The Analytical Development Section of SRTC was requested by the Facilities Disposition Division (FDD) to determine the holdup of enriched uranium in the 321-M facility as part of an overall deactivation project of the facility. The 321-M facility was used to fabricate enriched uranium fuel assemblies, lithium-aluminum target tubes, neptunium assemblies, and miscellaneous components for the production reactors. The facility also includes the 324-M storage building and the passageway connecting it to 321-M. The results of the holdup assays are essential for determining compliance with the Solid Waste's Waste Acceptance Criteria, Material Control and Accountability, and to meet criticality safety controls. Two measurement systems will be used to determine highly enriched uranium (HEU) holdup: One is a portable HPGe detector and EG and G Dart system that contains high voltage power supply and signal processing electronics. A personal computer with Gamma-Vision software was used to provide an MCA card, and space to store and manipulate multiple 4096-channel g-ray spectra. The other is a 2 inches x 2 inches NaI crystal with an MCA that uses a portable computer with a Canberra NaI plus card installed. This card converts the PC to a full function MCA and contains the ancillary electronics, high voltage power supply and amplifier, required for data acquisition. This report describes and documents the HPGe point, line, area, and constant geometry-constant transmission detector efficiency calibrations acquired and calculated for use in conducting holdup measurements as part of the overall deactivation project of building 321-M.

  8. Standard guide for calibrating reticles and light microscope magnifications

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2002-01-01

    1.1 This guide covers methods for calculating and calibrating microscope magnifications, photographic magnifications, video monitor magnifications, grain size comparison reticles, and other measuring reticles. Reflected light microscopes are used to characterize material microstructures. Many materials engineering decisions may be based on qualitative and quantitative analyses of a microstructure. It is essential that microscope magnifications and reticle dimensions be accurate. 1.2 The calibration using these methods is only as precise as the measuring devices used. It is recommended that the stage micrometer or scale used in the calibration should be traceable to the National Institute of Standards and Technology (NIST) or a similar organization. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory lim...

  9. Development of Hydrocarbon Flow Calibration Facility as a National Standard

    Science.gov (United States)

    Shimada, Takashi; Doihara, Ryouji; Terao, Yoshiya; Takamoto, Masaki

    A new primary standard for hydrocarbon flow measurements has been constructed at National Metrology Institute of Japan (NMIJ). The facility was designed for the calibration of hydrocarbon flowmeters in the flow rate range between 3 and 300 m3/h. The expanded uncertainty is estimated to be 0.03 % for volumetric flow rate and 0.02 % for mass flow rate (coverage factor: k = 2). The primary standard is based on a static and gravimetric method with a flying start and finish. The facility consists of two test rigs using kerosene and light oil as working fluids. The test lines for the flowmeters are 50, 100 and 150 mm in diameter and three servo positive displacement meters are used as working standards. To verify the calibration performance, a Coriolis flowmeter, a turbine meter and a positive displacement flowmeter have been calibrated at both test rigs. Furthermore, an international comparison with SP, Swedish National Testing Research Institute, was carried out. A screw-type positive displacement flowmeter was selected as the transfer standard and was calibrated at NMIJ and SP. The result shows that the two national standards at the two institutes agree within the quoted expanded uncertainties.

  10. Standard practice for torque calibration of testing machines and devices

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2009-01-01

    1.1 This practice covers procedures and requirements for the calibration of torque for static and quasi-static torque capable testing machines or devices. These may, or may not, have torque indicating systems and include those devices used for the calibration of hand torque tools. Testing machines may be calibrated by one of the three following methods or combination thereof: 1.1.1 Use of standard weights and lever arms. 1.1.2 Use of elastic torque measuring devices. 1.1.3 Use of elastic force measuring devices and lever arms. 1.1.4 Any of the methods require a specific uncertainty of measurement and a traceability derived from national standards of mass and length. 1.2 The procedures of 1.1.1, 1.1.2, and 1.1.3 apply to the calibration of the torque-indicating systems associated with the testing machine, such as a scale, dial, marked or unmarked recorder chart, digital display, etc. In all cases the buyer/owner/user must designate the torque-indicating system(s) to be calibrated and included in the repor...

  11. On Calibrations Using the Crab Nebula as a Standard Candle

    Science.gov (United States)

    Weisskopf, Martin; Guainazzi, Matteo; Jahoda, Keith; Shaposhnikov, Nikolai; ODell, Stephen; Zavlin, Vyacheslav; Wilson-Hodge, Colleen; Elsner, Ronald

    2009-01-01

    Inspired by a recent paper (Kirsch et al. 2005) on possible use of the Crab Nebula as a standard candle for calibrating X-ray response func tions, we examine possible consequences of intrinsic departures from a single (absorbed) power law upon such calibrations. We limited our analyses to three more modern X-ray instruments -- the ROSAT/PSPC, th e RXTE/PCA, and the XMM-Newton/EPIC-pn. The results are unexpected an d indicate a need to refine two of the three response functions studi ed. The implications for Chandra will be discussed.

  12. Monochromator-Based Absolute Calibration of a Standard Radiation Thermometer

    Science.gov (United States)

    Mantilla, J. M.; Hernanz, M. L.; Campos, J.; Martín, M. J.; Pons, A.; del Campo, D.

    2014-04-01

    Centro Español de Metrología (CEM) is disseminating the International Temperature Scale (ITS-90), at high temperatures, by using the fixed points of Ag and Cu and a standard radiation thermometer. However, the future mise-en-pratique for the definition of the kelvin ( MeP-K) will include the dissemination of the kelvin by primary methods and by indirect approximations capable of exceptionally low uncertainties or increased reliability. Primary radiometry is, at present, able to achieve uncertainties competitive with the ITS-90 above the silver point with one of the possible techniques the calibration for radiance responsivity of an imaging radiometer (radiance method). In order to carry out this calibration, IO-CSIC (Spanish Designated Institute for luminous intensity and luminous flux) has collaborated with CEM, allowing traceability to its cryogenic radiometer. A monochromator integrating sphere-based spectral comparator facility has been used to calibrate one of the CEM standard radiation thermometers. The absolute calibrated standard radiation thermometer has been used to determine the temperatures of the fixed points of Cu, Co-C, Pt-C, and Re-C. The results obtained are 1357.80 K, 1597.10 K, 2011.66 K, and 2747.64 K, respectively, with uncertainties ranging from 0.4 K to 1.1 K.

  13. Calibration of the Standards and Calibration Laboratory`s Co{sup 60} Radiation Pool

    Energy Technology Data Exchange (ETDEWEB)

    Wirtenson, G.R.; White, R.H.

    1993-01-01

    The authors report measurements of dose rates at various locations in the LLNL Standards and Calibrations Laboratory`s Co{sup 60} Radiation Pool. Plots show the dependence of dose rate on radius near the bottom of the pool and the dependence of dose rate on height at a fixed distance from the pool center. The effect of varying sample location within the pool`s dry-well was also investigated.

  14. Development of a low-level Ar-37 calibration standard

    CERN Document Server

    Williams, R M; Bowyer, T W; Day, A R; Fuller, E S; Haas, D A; Hayes, J C; Hoppe, E W; Humble, P H; Keillor, M E; LaFerriere, B D; Mace, E K; McIntyre, J I; Miley, H S; Myers, A W; Orrell, J L; Overman, C T; Panisko, M E; Seifert, A

    2016-01-01

    Argon-37 is an environmental signature of an underground nuclear explosion. Producing and quantifying low-level Ar-37 standards is an important step in the development of sensitive field measurement instruments. This paper describes progress at Pacific Northwest National Laboratory in developing a process to generate and quantify low-level Ar-37 standards, which can be used to calibrate sensitive field systems at activities consistent with soil background levels. This paper presents a discussion of the measurement analysis, along with assumptions and uncertainty estimates.

  15. Performance of calibration standards for antigen quantitation with flow cytometry.

    Science.gov (United States)

    Lenkei, R; Gratama, J W; Rothe, G; Schmitz, G; D'hautcourt, J L; Arekrans, A; Mandy, F; Marti, G

    1998-10-01

    In the frame of the activities initiated by the Task Force for Antigen Quantitation of the European Working Group on Clinical Cell Analysis (EWGCCA), an experiment was conducted to evaluate microbead standards used for quantitative flow cytometry (QFCM). An unified window of analysis (UWA) was established on three different instruments (EPICS XL [Coulter Corporation, Miami, FL], FACScan and FACS Calibur [Becton Dickinson, San Jose, CA]) with QC3 microbeads (FCSC, PR). By using this defined fluorescence intensity scale, the performance of several monoclonal antibodies directed to CD3, CD4, and CD8 (conjugated and unconjugated), from three manufacturers (BDIS, Coulter [Immunotech], and DAKO) was tested. In addition, the QIFI system (DAKO) and QuantiBRITE (BDIS), and a method of relative fluorescence intensity (RFI, method of Giorgi), were compared. mAbs reacting with three more antigens, CD16, CD19, and CD38 were tested on the FACScan instrument. Quantitation was carried out using a single batch of cryopreserved peripheral blood leukocytes, and all tests were performed as single color analyses. Significant correlations were observed between the antibody-binding capacity (ABC) values of the same CD antigen measured with various calibrators and with antibodies differing in respect to vendor, labeling and possible epitope recognition. Despite the significant correlations, the ABC values of most monoclonal antibodies differed by 20-40% when determined by the different fluorochrome conjugates and different calibrators. The results of this study indicate that, at the present stage of QFCM consistent ABC values may be attained between laboratories provided that a specific calibration system is used including specific calibrators, reagents, and protocols.

  16. A report from the AVS Standards Committee - Comparison of ion gauge calibrations by several standards laboratories

    Science.gov (United States)

    Warshawsky, I.

    1982-01-01

    Calibrations by four U.S. laboratories of four hot-cathode ion gauges, in the range 0.07-13 mPa, showed systematic differences among laboratories that were much larger than the expected error of any one calibration. They also suggested that any of the four gauges tested, if properly packaged and shipped, was able to serve as a transfer standard with probable error of 2%. A second comparison was made of the calibrations by two U.S. laboratories of some other gauges that had also been calibrated by the National Physical Laboratory, England. Results did not permit conclusive determination of whether differences were due to the laboratories or to changes in the gauges.

  17. Internal standard versus external standard calibration: an uncertainty case study of a liquid chromatography analysis

    Directory of Open Access Journals (Sweden)

    Elcio Cruz de Oliveira

    2010-01-01

    Full Text Available Traditionally, in the cigarettes industry, the determination of ammonium ion in the mainstream smoke is performed by ion chromatography. This work studies this determination and compares the results of this technique with the use of external and internal standard calibration. A reference cigarette sample presented measurement uncertainty of 2.0 μg/cigarette and 1.5 μg/cigarette, with external and internal standard, respectively. It is observed that the greatest source of uncertainty is the bias correction factor and that it is even more significant when using external standard, confirming thus the importance of internal standardization for this correction.

  18. Electronic Correlated Noise Calibration Standard for Interferometric and Polarimetric Microwave Radiometers Project

    Data.gov (United States)

    National Aeronautics and Space Administration — A new type of calibration standard is proposed which produces a pair of microwave noise signals to aid in the characterization and calibration of correlating...

  19. Accuracy, standardization, and interlaboratory calibration standards for foraminiferal Mg/Ca thermometry

    Science.gov (United States)

    Greaves, Mervyn; Barker, Stephen; Daunt, Caroline; Elderfield, Henry

    2005-02-01

    The use of liquid and solid standards for foraminiferal Mg/Ca and Sr/Ca determinations and interlaboratory calibration has been investigated. Preparation of single element standard solutions from primary solid standard material enables the preparation of mixed standard solutions with Mg/Ca and Sr/Ca ratios of known accuracy to better than 0.1%. We also investigated commercial reference materials to determine whether existing carbonate standards could be used as reference material for Mg/Ca determinations in foraminiferal calcite. We propose that, in the absence of a pure calcium carbonate standard certified for Mg/Ca, ECRM 752-1, a limestone CRM containing Mg/Ca within the range of typical foraminifera, is a suitable solid standard for interlaboratory calibration. Replicate Mg/Ca determinations showed that, provided silicate phases are removed by centrifugation, this material is homogenous within the precision of daily instrumental Mg/Ca determinations over a range of sample weights from 10 to 1000 mg, taken from two separate bottles of ECRM 752-1. Results gave an average value of Mg/Ca = 3.75 mmol/mol (0.015 s.d., 0.41% r.s.d.) on 118 determinations from the two bottles.

  20. The Moon as a photometric calibration standard for microwave sensors

    Science.gov (United States)

    Burgdorf, Martin; Buehler, Stefan A.; Lang, Theresa; Michel, Simon; Hans, Imke

    2016-08-01

    Instruments on satellites for Earth observation on polar orbits usually employ a two-point calibration technique, in which deep space and an onboard calibration target provide two reference flux levels. As the direction of the deep-space view is in general close to the celestial equator, the Moon sometimes moves through the field of view and introduces an unwelcome additional signal. One can take advantage of this intrusion, however, by using the Moon as a third flux standard, and this has actually been done for checking the lifetime stability of sensors operating at visible wavelengths. As the disk-integrated thermal emission of the Moon is less well known than its reflected sunlight, this concept can in the microwave range only be used for stability checks and intercalibration. An estimate of the frequency of appearances of the Moon in the deep-space view, a description of the limiting factors of the measurement accuracy and models of the Moon's brightness, and a discussion of the benefits from complementing the naturally occurring appearances of the Moon with dedicated spacecraft maneuvers show that it would be possible to detect photometric lifetime drifts of a few percent with just two measurements. The pointing accuracy is the most crucial factor for the value of this method. Planning such observations in advance would be particularly beneficial, because it allows observing the Moon at well-defined phase angles and putting it at the center of the field of view. A constant phase angle eliminates the need for a model of the Moon's brightness when checking the stability of an instrument. With increasing spatial resolution of future microwave sensors another question arises, viz. to what extent foreground emission from objects other than the Moon will contaminate the flux entering the deep-space view, which is supposed to originate exclusively in the cosmic microwave background. We conclude that even the brightest discreet sources have flux densities below the

  1. On the long-term stability of calibration standards in different matrices.

    Science.gov (United States)

    Kandić, A; Vukanac, I; Djurašević, M; Novković, D; Šešlak, B; Milošević, Z

    2012-09-01

    In order to assure Quality Control in accordance with ISO/IEC 17025, it was important, from metrological point of view, to examine the long-term stability of calibration standards previously prepared. Comprehensive reconsideration on efficiency curves with respect to the ageing of calibration standards is presented in this paper. The calibration standards were re-used after a period of 5 years and analysis of the results showed discrepancies in efficiency values. PMID:22405642

  2. Standard practice for alternate actinide calibration for inductively coupled plasma-mass spectrometry

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2004-01-01

    1.1 This practice provides guidance for an alternate linear calibration for the determination of selected actinide isotopes in appropriately prepared aqueous solutions by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS). This alternate calibration is mass bias adjusted using thorium-232 (232Th) and uranium-238 (238U) standards. One of the benefits of this standard practice is the ability to calibrate for the analysis of highly radioactive actinides using calibration standards at much lower specific activities. Environmental laboratories may find this standard practice useful if facilities are not available to handle the highly radioactive standards of the individual actinides of interest. 1.2 The instrument response for a series of determinations of known concentration of 232Th and 238U defines the mass versus response relationship. For each standard concentration, the slope of the line defined by 232Th and 238U is used to derive linear calibration curves for each mass of interest using interference equ...

  3. Development of a Low-Level Ar-37 Calibration Standard

    Energy Technology Data Exchange (ETDEWEB)

    Williams, Richard M.; Aalseth, Craig E.; Bowyer, Ted W.; Day, Anthony R.; Fuller, Erin S.; Haas, Derek A.; Hayes, James C.; Hoppe, Eric W.; Humble, Paul H.; Keillor, Martin E.; LaFerriere, Brian D.; Mace, Emily K.; McIntyre, Justin I.; Miley, Harry S.; Myers, Allan W.; Orrell, John L.; Overman, Cory T.; Panisko, Mark E.; Seifert, Allen

    2016-03-07

    Argon-37 is an important environmental signature of an underground nuclear explosion. Producing and quantifying low-level 37Ar standards is an important step in the development of sensitive field measurement instruments for use during an On-Site Inspection, a key provision of the Comprehensive Nuclear-Test-Ban Treaty. This paper describes progress at Pacific Northwest National Laboratory (PNNL) in the development of a process to generate and quantify low-level 37Ar standard material, which can then be used to calibrate sensitive field systems at activities consistent with soil background levels. The 37Ar used for our work was generated using a laboratory-scale, high-energy neutron source to irradiate powdered samples of calcium carbonate. Small aliquots of 37Ar were then extracted from the head space of the irradiated samples. The specific activity of the head space samples, mixed with P10 (90% stable argon:10% methane by mole fraction) count gas, is then derived using the accepted Length-Compensated Internal-Source Proportional Counting method. Due to the low activity of the samples, a set of three Ultra-Low Background Proportional-Counters designed and fabricated at PNNL from radio-pure electroformed copper was used to make the measurements in PNNL’s shallow underground counting laboratory. Very low background levels (<10 counts/day) have been observed in the spectral region near the 37Ar emission feature at 2.8 keV. Two separate samples from the same irradiation were measured. The first sample was counted for 12 days beginning 28 days after irradiation, the second sample was counted for 24 days beginning 70 days after irradiation (the half-life of 37Ar is 35.0 days). Both sets of measurements were analyzed and yielded very similar results for the starting activity (~0.1 Bq) and activity concentration (0.15 mBq/ccSTP argon) after P10 count gas was added. A detailed uncertainty model was developed based on the ISO Guide to the Expression of Uncertainty in

  4. Calibration Standards for Surface Topography Measuring Systems down to Nanometric Range

    DEFF Research Database (Denmark)

    Trumpold, H.; De Chiffre, Leonardo; Andreasen, Jan Lasson;

    Background For the precise and accurate measurement of surface topography a whole range of surface detection systems is available. With their application in research and production problems arise due to the lack of traceable standard artefacts for the instrument calibration in X, Y and Z directions...... and for the calibration of filters. Existing ISO standards on calibration specimens are inadequate and limited in that they only cover contacting instruments and only partially the measuring ranges for these instruments. The whole range of non-contacting instruments are not covered despite their increasing use...... in industries ranging from automotive manufacture to ultraprecision manufacture of data storage systems and compact discs. Objectives The project is concerned with developing calibration standards including their production methods and calibration procedures as a consistent means of calibrating different types...

  5. Lyophilized standards for the calibration of real time PCR assay for hepatitis C virus RNA

    Institute of Scientific and Technical Information of China (English)

    WANG Lu-nan; WU Jian-min; DENG Wei; SHEN Zi-yu; CHEN Wen-xiang; LI Jin-ming

    2006-01-01

    Background Since October 1997, an international standard for hepatitis C virus (HCV) nucleic acid amplification technology assay, 96/790, has been available. We compared a series of lyophilized standards with known HCV RNA concentrations against the international standard in fluorescence quantitative PCR detection.Methods A series of lyophilized sera were calibrated by ROCHE COBAS AMPLICOR HCV Monitor test against the international standard and sent to various manufacturers to analyse the samples using their own kits.Then calibration curves from the series were compared with that obtained from the external standard calibration curve with the manufacture's series.Results The standard calibration curve with the series of lyophilized serum showed an excellent correlation(R2>0.98), slope and intercept that were similar to those from the manufacture's series. When the standard calibration curve from the series of lyophilized standards were used to define the values of the given sample,lower coefficients of variation between kits from different manufactures were obtained.Conclusion The results showed that the lyophilized standards could be used to setup the standard calibration curve for clinical HCV RNA quantitative PCR detection.

  6. Source geometry factors for HDR 192Ir brachytherapy secondary standard well-type ionization chamber calibrations

    Science.gov (United States)

    Shipley, D. R.; Sander, T.; Nutbrown, R. F.

    2015-03-01

    Well-type ionization chambers are used for measuring the source strength of radioactive brachytherapy sources before clinical use. Initially, the well chambers are calibrated against a suitable national standard. For high dose rate (HDR) 192Ir, this calibration is usually a two-step process. Firstly, the calibration source is traceably calibrated against an air kerma primary standard in terms of either reference air kerma rate or air kerma strength. The calibrated 192Ir source is then used to calibrate the secondary standard well-type ionization chamber. Calibration laboratories are usually only equipped with one type of HDR 192Ir source. If the clinical source type is different from that used for the calibration of the well chamber at the standards laboratory, a source geometry factor, ksg, is required to correct the calibration coefficient for any change of the well chamber response due to geometric differences between the sources. In this work we present source geometry factors for six different HDR 192Ir brachytherapy sources which have been determined using Monte Carlo techniques for a specific ionization chamber, the Standard Imaging HDR 1000 Plus well chamber with a type 70010 HDR iridium source holder. The calculated correction factors were normalized to the old and new type of calibration source used at the National Physical Laboratory. With the old Nucletron microSelectron-v1 (classic) HDR 192Ir calibration source, ksg was found to be in the range 0.983 to 0.999 and with the new Isodose Control HDR 192Ir Flexisource ksg was found to be in the range 0.987 to 1.004 with a relative uncertainty of 0.4% (k = 2). Source geometry factors for different combinations of calibration sources, clinical sources, well chambers and associated source holders, can be calculated with the formalism discussed in this paper.

  7. Legacy of NOAA, NASA and NIST Cooperation in Developing Radiometric Calibration Standards, Equipment and Methodologies

    OpenAIRE

    Datla, Raju; Weinreb, Michael; Cao, Changyong

    2012-01-01

    This report traces the historical development of radiometric standards and calibration methodologies over the past several decades for optical passive sensor calibration at the National Oceanic and Atmospheric Administration (NOAA), National Aeronautic and Space Administration (NASA) and National Institute of Standards and Technology (NIST), formerly National Bureau of Standards NBS. From the beginning, a remarkable goal oriented interaction between scientists and engineers of the 3 agencies ...

  8. Characterization of a 137Cs standard source for calibration purposes ar CRCN-NE

    International Nuclear Information System (INIS)

    Radiation protection monitoring instruments should be calibrated by accredited calibration laboratories. To offer calibration services, a laboratory must accomplish all requirements established by the national regulatory agency. The Calibration Service of the Centro Regional de Ciencias Nucleares (CRCN-NE), Comissao Nacional de Energia Nuclear, Recife, Brazil, is trying to achieve this accreditation. In the present work, a 137Cs standard source was characterized following the national and international recommendations and the results are presented. This source is a commercially available single source irradiator model 28-8A, manufactured by JLShepherd and Associates, with initial activity of 444 GBq (05/13/03). To provide different air kerma rates, as required for the calibration of portable radiation monitors, this irradiator have a set of four lead attenuators with different thickness, providing attenuation factors equal to 2, 4, 10 and 100 times (nominally). The performed tests included: size and uniformity of the radiation standard field at calibration reference position, variation of the air kerma rate for different lead attenuators, determination of attenuation factors for each lead attenuator configuration, and determination of the radiation scattering at the calibration reference position. The results showed the usefulness of the 137Cs standard source for the calibration of radiation protection monitoring detectors. (author)

  9. Services of the CDRH X-ray calibration laboratory and their traceability to National Standards

    Energy Technology Data Exchange (ETDEWEB)

    Cerra, F.; Heaton, H.T. [Center for Devices and Radiological Health, Rockville, MD (United States)

    1993-12-31

    The X-ray Calibration Laboratory (XCL) of the Center for Devices and Radiological Health (CDRH) provides calibration services for the Food and Drug Administration (FDA). The instruments calibrated are used by FDA and contract state inspectors to verify compliance with federal x-ray performance standards and for national surveys of x-ray trends. In order to provide traceability of measurements, the CDRH XCL is accredited by the National Voluntary Laboratory Accreditation Program (NVLAP) for reference, diagnostic, and x-ray survey instrument calibrations. In addition to these accredited services, the CDRH XCL also calibrates non-invasive kVp meters in single- and three-phase x-ray beams, and thermoluminescent dosimeter (TLD) chips used to measure CT beam profiles. The poster illustrates these services and shows the traceability links back to the National Standards.

  10. Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry

    International Nuclear Information System (INIS)

    Purpose: The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by ∼25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. Methods: The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Results: Calibration precision was measured to be better than 5%–7%, 3%–5%, and 2%–4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.7 ± 1.1 mV cGy−1 versus the CT scatter phantom 29.2 ± 1.0 mV cGy−1 and FIA with x-ray 29.9 ± 1.1 mV cGy−1 methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of ∼3000 mV. Conclusions: The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration coefficients for the eventual

  11. Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Brady, S. L.; Kaufman, R. A. [Department of Radiological Sciences, St. Jude Children' s Research Hospital, Memphis, Tennessee 38105 (United States)

    2012-06-15

    Purpose: The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by {approx}25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. Methods: The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Results: Calibration precision was measured to be better than 5%-7%, 3%-5%, and 2%-4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.7 {+-} 1.1 mV cGy{sup -1} versus the CT scatter phantom 29.2 {+-} 1.0 mV cGy{sup -1} and FIA with x-ray 29.9 {+-} 1.1 mV cGy{sup -1} methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of {approx}3000 mV. Conclusions: The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration

  12. Establishment of the platinum-iridium kilogram mass standards at NMIJ after the Extraordinary Calibrations

    Science.gov (United States)

    Mizushima, Shigeki; Fujii, Kenichi

    2016-04-01

    The Bureau International des Poids et Mesures has carried out calibrations of the platinum-iridium kilogram mass standards by referencing the international prototype of the kilogram for the first time since the third periodic verification of national prototypes of the kilogram was carried out in 1988-92. This calibration campaign was designated ‘Extraordinary Calibrations’ in the second phase, in which two platinum-iridium kilogram mass standards of the National Metrology Institute of Japan were calibrated with a standard uncertainty of 3.5 μg. By adding these new calibration data into our data sets from 1991, we established our mass unit with a standard uncertainty of 3.3 μg by least-squares analysis using an exponential model, which is useful for compensating for mass increase after cleaning the mass standards. Moreover, it was found that our established mass unit following the Extraordinary Calibrations shifted against our previously maintained mass unit by  -20.8 μg as of the beginning of 2015. The analysis with a linear model revealed that the amount of mass increase over time of some standards was significantly smaller than that suggested at the third periodic verification of national prototypes of the kilogram. The analysis with the exponential model gave an exponent of 0.217 with a standard uncertainty of 0.057. This suggests that the mass increase due to surface contamination cannot be explained by a diffusion-limited process.

  13. Standardization of Laser Methods and Techniques for Vibration Measurements and Calibrations

    Science.gov (United States)

    von Martens, Hans-Jürgen

    2010-05-01

    The realization and dissemination of the SI units of motion quantities (vibration and shock) have been based on laser interferometer methods specified in international documentary standards. New and refined laser methods and techniques developed by national metrology institutes and by leading manufacturers in the past two decades have been swiftly specified as standard methods for inclusion into in the series ISO 16063 of international documentary standards. A survey of ISO Standards for the calibration of vibration and shock transducers demonstrates the extended ranges and improved accuracy (measurement uncertainty) of laser methods and techniques for vibration and shock measurements and calibrations. The first standard for the calibration of laser vibrometers by laser interferometry or by a reference accelerometer calibrated by laser interferometry (ISO 16063-41) is on the stage of a Draft International Standard (DIS) and may be issued by the end of 2010. The standard methods with refined techniques proved to achieve wider measurement ranges and smaller measurement uncertainties than that specified in the ISO Standards. The applicability of different standardized interferometer methods to vibrations at high frequencies was recently demonstrated up to 347 kHz (acceleration amplitudes up to 350 km/s2). The relative deviations between the amplitude measurement results of the different interferometer methods that were applied simultaneously, differed by less than 1% in all cases.

  14. Advisory Committee for the Calibration Standards of Ionizing Radiation Measurement

    International Nuclear Information System (INIS)

    An account of the activity during the past two years and of the plans for future work is given for the three Sections of the Comite Consultatif pour les Etalons de Mesure des Rayonnements Ionisants. Section I (Rayons X et #betta#, electrons) studied in detail the results of an intercomparison of Frike dosimeters. A recommendation was made concerning the possibility of expressing calibrations made in terms of exposure in terms of air kerma or water kerma. Section II (Mesure des radionucleides) studied the results of recent international comparisons (55Fe, 133Ba and 134Cs) and made plans for new ones. Section III (Mesures neutroniques) presented the status of the international comparisons of neutron fluence rate in progress and decided to organize new ones. The reports of the Section chairmen are followed by the presentation of the work carried out at BIPM by the corresponding groups. The status of the proposal by Section III for a 14 MeV neutron dosimetry facility at BIPM is discussed in detail and a proposal is made for a neutron dosimetry intercomparison

  15. Calibration of the SH134-20 Standard Gain Horn

    DEFF Research Database (Denmark)

    Pivnenko, Sergey; Breinbjerg, Olav

    This report documents the measurement of the linearly polarized SH134-20 Standard Gain Horn. The measurement comprises on-axis gain, on-axis polarization characteristics, and reflection coefficient at 111 frequencies in the frequency range from 22-33 GHz. The measurement was carried out at the DT......-ESA Spherical Near-Field Antenna Test Facility in December 2012 for SE Laboratories, Santa Clara, CA, USA....

  16. Calibration of beta-particle ophthalmic applicators at the National Bureau of Standards

    International Nuclear Information System (INIS)

    The method used at the National Bureau of Standards for the calibration of strontium-90 + yttrium-90 beta-particle ophthalmic applicators in terms of absorbed dose to water, is described. The method involves measurement of ionization density at the applicator surface with an extrapolation chamber, correction for the difference in backscatter between the collection electrode and water, and application of the Bragg-Gray equation. The calibration obtained is an average over the active surface of the applicator. The overall uncertainty of the surface calibration is about + or - 15 percent

  17. Calibration of survey meters at the Algerian Secondary Standard Dosimetry Laboratory

    International Nuclear Information System (INIS)

    Full text: The Algerian Secondary Standard Dosimetry Laboratory, which became a Regional AFRA Designated Centre for French spoken countries, in the field of calibration since 2005, has developed and implemented methodologies for the calibration of radiation protection instruments such as survey meters. These instruments were initially calibrated in terms of air kerma free in air (NK) in 60Co, 137Cs gamma radiations. Although the recommended calibration quantity is H*(10) (Sv/h), most of instruments measure different physical quantities and problems about the expression of calibration factor in the calibration certificates are always raised. In order to switch to calibrations in terms of H*(10), a feasibility study was performed using instruments submitted for calibration at the SSDL. This paper presents the variation of the calibration coefficients, Fc, obtained since 2005 for three types of survey meters. The uncertainty components involved in the determination of Fc are analysed in details. Material and method: This study included the most used survey meters in Algeria and received at the SSDL in 2005 and 2006 which consist of 127 survey-meters of type Automess (different models), 80 Ludlum model 3 and 23 Graetz X 5 DE. Calibrations were performed using the beam output decay method. This output was determined with the Reference Standard chamber of type NE 2575 in 2005 and LS01 in 2006, both calibrated at the IAEA calibration laboratory. The average calibration coefficients, Fc, which converts the instruments reading to ambient equivalent absorbed dose, are calculated and the relative standard deviations assessed. The variation of calibration coefficients, for each type of instruments, is compared to the overall stated uncertainty for this coefficient. Results and discussion: The mean calibration coefficients in terms of H*(10) were lying from 0.993 ± 2.55% and 1.039 ± 3.16% for the Automess survey meters and from . The results are 1.215 ± 5.93%, 1.149 ± 14

  18. Nanoscale Calibration Standards and Methods: Dimensional and Related Measurements in the Micro- and Nanometer Range

    Science.gov (United States)

    Wilkening, Günter; Koenders, Ludger

    2005-08-01

    The quantitative determination of the properties of micro- and nanostructures is essential in research and development. It is also a prerequisite in process control and quality assurance in industry. The knowledge of the geometrical dimensions of structures in most cases is the base, to which other physical and chemical properties are linked. Quantitative measurements require reliable and stable instruments, suitable measurement procedures as well as appropriate calibration artefacts and methods. The seminar "NanoScale 2004" (6th Seminar on Quantitative Microscopy and 2nd Seminar on Nanoscale Calibration Standards and Methods) at the National Metrology Institute (Physikalisch-Technische Bundesanstalt PTB), Braunschweig, Germany, continues the series of seminars on Quantitative Microscopy. The series stimulates the exchange of information between manufacturers of relevant hard- and software and the users in science and industry. Topics addressed in these proceedings are a) the application of quantitative measurements and measurement problems in: microelectronics, microsystems technology, nano/quantum/molecular electronics, chemistry, biology, medicine, environmental technology, materials science, surface processing b) calibration & correction methods: calibration methods, calibration standards, calibration procedures, traceable measurements, standardization, uncertainty of measurements c) instrumentation and methods: novel/improved instruments and methods, reproducible probe/sample positioning, position-measuring systems, novel/improved probe/detector systems, linearization methods, image processing

  19. Calibration of Modulation Transfer Function of Surface Profilometers with 1D and 2D Binary Pseudo-random Array Standards

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy V.; McKinney, Wayne R.; Takacs, Peter Z.

    2008-05-19

    We suggest and describe the use of a binary pseudo-random grating as a standard test surface for calibration of the modulation transfer function of microscopes. Results from calibration of a MicromapTM-570 interferometric microscope are presented.

  20. Investigation of Ultrasonic Calibration using Steel Standard Reference Blocks

    Directory of Open Access Journals (Sweden)

    Mohamed Galal Sayed ALI

    2012-12-01

    Full Text Available The equipment used in standardization must be adjusted in order that both the receiver amplifier and video display are linear over the range of distance and flaw size to be used. The failure to do that would result in different block-to-block response ratios because of the measurements depending on the scope reading or attenuator setting. In this paper, using two sets of steel reference blocks (area-amplitude and distance-amplitude to modified method to verify the linearity of the system of instruments used in the ultrasonic inspection. This method depends on the response from reference blocks with different hole diameters at constant gain and ratio between the figures slope at different gains. The linearity of the ultrasonic instruments can be measured with high precision. The results can be used effectively for determining resolution and sensitivity capabilities.

  1. Uncertainty Analysis of Spectral Irradiance Reference Standards Used for NREL Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Habte, A.; Andreas, A.; Reda, I.; Campanelli, M.; Stoffel, T.

    2013-05-01

    Spectral irradiance produced by lamp standards such as the National Institute of Standards and Technology (NIST) FEL-type tungsten halogen lamps are used to calibrate spectroradiometers at the National Renewable Energy Laboratory. Spectroradiometers are often used to characterize spectral irradiance of solar simulators, which in turn are used to characterize photovoltaic device performance, e.g., power output and spectral response. Therefore, quantifying the calibration uncertainty of spectroradiometers is critical to understanding photovoltaic system performance. In this study, we attempted to reproduce the NIST-reported input variables, including the calibration uncertainty in spectral irradiance for a standard NIST lamp, and quantify uncertainty for measurement setup at the Optical Metrology Laboratory at the National Renewable Energy Laboratory.

  2. Transfer standard device to improve the traceable calibration of physiotherapy ultrasound machines

    NARCIS (Netherlands)

    Hekkenberg, R.T.; Richards, A.; Beissner, K.; Zeqiri, B.; Bezemer, R.A.; Hodnett, M.; Prout, G.; Cantrall, C.

    2006-01-01

    Ultrasound (US) physiotherapy as a clinical treatment is extremely common in the Western world. Internationally, regulation to ensure safe application of US physiotherapy by regular calibration ranges from nil to mandatory. The need for a portable power standard (PPS) has been addressed within a Eur

  3. EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards (2012 Revision)

    Science.gov (United States)

    In 1997, the U.S. Environmental Protection Agency (EPA) in Research Triangle Park, North Carolina, revised its 1993 version of its traceability protocol for the assay and certification of compressed gas and permeation-device calibration standards. The protocol allows producers o...

  4. Secondary calibration laboratory for ionizing radiation laboratory accreitation program National Institute of Standards and Technology National Voluntary Laboratory Accreditation Program

    Energy Technology Data Exchange (ETDEWEB)

    Martin, P.R.

    1993-12-31

    This paper presents an overview of the procedures and requirements for accreditation under the Secondary Calibration Laboratory for Ionizing Radiation Program (SCLIR LAP). The requirements for a quality system, proficiency testing and the onsite assessment are discussed. The purpose of the accreditation program is to establish a network of secondary calibration laboratories that can provide calibrations traceable to the primary national standards.

  5. SIMULTANEOUS CALIBRATION OF MOLECULAR WEIGHT SEPARATION AND COLUMN DISPERSION OF SEC WITH CHARACTERIZED POLYMER STANDARDS

    Institute of Scientific and Technical Information of China (English)

    CHENG Rongshi; BO Shuqin

    1983-01-01

    With the aid of the theoretical relationship between the calibration relation of a SEC column for the monodisperse polymer species under ideal working condition and the effective relations between the molecular weight and the elution volume for characterized polymer samples, a computational procedure for simultaneous calibration of molecular weight separation and column dispersion is proposed. From the experimental chromatograms of narrow MWD polystyrene standards and broad MWD 1,2-polybutadiene fractions the spreading factors of a SEC column was deduced by the proposed method. The variation of the spreading factor with the elution volume is independent upon the polymer sample used.

  6. A wideband calibration procedure for symmetric planar sensors using an unknown load standard

    International Nuclear Information System (INIS)

    A new calibration method, suitable for planar sensors and other non-coaxial devices, is presented. Assuming that the matching network of the device under test is symmetrical, simplifications are made, leading to a procedure that uses a single parallel unknown standard named TYS (thru-Y-side). The de-embedding equations are developed and the results of its application are reported, showing a bandwidth wider than thru-reflect-match. Finally, the calibration is applied to the characterization of a planar sensor and the permittivity determination of three known materials, compared with the widely used thru-reflect-line method. (paper)

  7. One-Port Direct/Reverse Method for Characterizing VNA Calibration Standards

    CERN Document Server

    Monsalve, Raul A; Mozdzen, Thomas J; Bowman, Judd D

    2016-01-01

    This paper introduces a one-port method for estimating model parameters of VNA calibration standards. The method involves measuring the standards through an asymmetrical passive network connected in direct mode and then in reverse mode, and using these measurements to compute the S-parameters of the network. The free parameters of the calibration standards are estimated by minimizing a figure of merit based on the expected equality of the S-parameters of the network when used in direct and reverse modes. The capabilities of the method are demonstrated through simulations, and real measurements are used to estimate the actual offset delay of a 50-$\\mathbf{\\Omega}$ calibration load that is assigned zero delay by the manufacturer. The estimated delay is $38.8$ ps with a $1\\sigma$ uncertainty of $2.1$ ps for this particular load. This result is verified through measurements of a terminated airline. The measurements agree better with theoretical models of the airline when the reference plane is calibrated using th...

  8. Development of a Primary Standard for Calibration of [18F]FDG Activity Measurement Systems

    International Nuclear Information System (INIS)

    The 18F national primary standard was developed by the INMRI-ENEA using the 4πβ Liquid Scintillation Spectrometry Method with 3H-Standard Efficiency Tracing. Measurements were performed at JRCIspra under a scientific collaboration between the Institute for Health and Consumer Production, the Amersham Health and the National Institute for Occupational Safety and Prevention (ISPESL). The goal of the work was to calibrate, with minimum uncertainty, the INMRI-ENEA transfer standard portable well-type ionisation chamber as well as other JRC-Ispra and Amersham Health reference Ionising Chambers used for FDG activity measurement

  9. X radiation qualities characterization following the standard IEC 61267 recommendations at the calibration laboratory of IPEN

    International Nuclear Information System (INIS)

    This work presents a methodology for the X radiation qualities characterization following the new recommendations of the standard 61267 of the International Electrotechnical Commission (IEC) to establish a new procedure for calibration of dosimetric systems used in the field of diagnostic radiology. The reference qualities radiation of IEC 61267: RQR 2 to RQR 10, RQA 2 to RQA 10, RQB 2 to RQB 10 and RQN 2 to RQN 10 were implanted at the calibration laboratory of IPEN (LCI). Their characteristics were analyzed through measurements of beam parameters such as: Practical peak voltage (PPV), specific additional filtrations for each qualities (high purity aluminum of about 99.9%), 1st and 2nd Half Value Layers, homogeneity coefficient. The inherent filtration of the X ray tube was also determined. With the establishment of these radiation qualities, the LCI will be ready to calibrate the measuring instruments of radiation in the new qualities, allowing an improvement in radiological services offered by IPEN. (author)

  10. A practical implementation of microphone free-field comparison calibration according to the standard IEC 61094-8

    DEFF Research Database (Denmark)

    Barrera Figueroa, Salvador; Torras Rosell, Antoni; Rasmussen, Knud;

    2012-01-01

    An international standard concerned with the calibration of microphones in a free field by comparison has recently been published. The standard contemplates two main calibration methodologies for determining the sensitivity of a microphone under test when compared against a reference microphone...

  11. Numerical Simulation and Experimental Validation of Calibrant-Loaded Extraction Phase Standardization Approach.

    Science.gov (United States)

    Alam, Md Nazmul; Pawliszyn, Janusz

    2016-09-01

    We present the kinetics of calibrant release and analyte uptake between the sample and calibrant-loaded extraction phase, CL-EP, with a finite-element analysis (FEA) using COMSOL Multiphysics software package. Effect of finite and infinite sample volume conditions, as well as various sample environment parameters such as fluid flow velocity, temperature, and presence of a binding matrix component were investigated in detail with the model in relation to the performance of the calibration. The simulation results supported by experimental data demonstrate the suitability of the CL-EP method for analysis of samples with variation of the sample environment parameters. The calibrant-loaded approach can provide both total and free concentrations from a single experiment based on whether the partition coefficient (Kes) value being used is measured in a matrix-matched sample or in a matrix-free sample, respectively. Total concentrations can also be obtained by utilizing CL-EP in combination with external matrix-matched calibrations, which can be employed to automate the sampling process and provide corrections for variations in sample preparation, matrix effects, and detection processes. This approach is also suitable for very small volumes of sample, where addition of an internal standard in the sample is either troublesome or can change the sample characteristics. PMID:27508421

  12. Standard test method for calibration of surface/stress measuring devices

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    1997-01-01

    Return to Contents page 1.1 This test method covers calibration or verification of calibration, or both, of surface-stress measuring devices used to measure stress in annealed and heat-strengthened or tempered glass using polariscopic or refractometry based principles. 1.2 This test method is nondestructive. 1.3 This test method uses transmitted light, and therefore, is applicable to light-transmitting glasses. 1.4 This test method is not applicable to chemically tempered glass. 1.5 Using the procedure described, surface stresses can be measured only on the “tin” side of float glass. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  13. Air kerma standardization for diagnostic radiology, and requirements proposal for calibration laboratories

    International Nuclear Information System (INIS)

    The demand for calibration services and quality control in diagnostic radiology has grown in the country since the publication of the governmental regulation 453, issued by the Ministry of Health in 1998. At that time, to produce results facing the new legislation, many laboratories used different standards and radiation qualities, some of which could be inadequate. The international standards neither supplied consistent radiation qualities and standardization for the different types of equipment available. This situation changed with the publication of the new edition of the IEC 61267 standard, published in 2005. A metrology network was created, but it is not yet accredited by the accreditation organism of the country, INMETRO. The objective of this work was to implement the standardization of the air kerma for the un attenuated qualities (RQR) of IEC 61267, and to develop a requirement proposal for instruments calibration laboratories. Results of interlaboratory comparisons demonstrate that the quantity is standardized and internationally traceable. A laboratory requirement proposal was finalized and it shall be submitted to INMETRO to be used as auxiliary normative document in laboratory accreditation. (author)

  14. PAS-cal: a Generic Recombinant Peptide Calibration Standard for Mass Spectrometry

    OpenAIRE

    Breibeck, Joscha; Serafin, Adam; Reichert, Andreas; Maier, Stefan; Küster, Bernhard; Skerra, Arne

    2014-01-01

    We describe the design, preparation, and mass-spectrometric characterization of a new recombinant peptide calibration standard with uniform biophysical and ionization characteristics for mass spectrometry. “PAS-cal” is an artificial polypeptide concatamer of peptide cassettes with varying lengths, each composed of the three small, chemically stable amino acids Pro, Ala, and Ser, which are interspersed by Arg residues to allow site-specific cleavage with trypsin. PAS-cal is expressed at high y...

  15. QconCAT standard for calibration of ion mobility-mass spectrometry systems.

    Science.gov (United States)

    Chawner, Ross; McCullough, Bryan; Giles, Kevin; Barran, Perdita E; Gaskell, Simon J; Eyers, Claire E

    2012-11-01

    Ion mobility-mass spectrometry (IM-MS) is a useful technique for determining information about analyte ion conformation in addition to mass/charge ratio. The physical principles that govern the mobility of an ion through a gas in the presence of a uniform electric field are well understood, enabling rotationally averaged collision cross sections (Ω) to be directly calculated from measured drift times under well-defined experimental conditions. However, such "first principle" calculations are not straightforward for Traveling Wave (T-Wave) mobility separations due to the range of factors that influence ion motion through the mobility cell. If collision cross section information is required from T-Wave mobility separations, then calibration of the instruments using known standards is essential for each set of experimental conditions. To facilitate such calibration, we have designed and generated an artificial protein based on the QconCAT technology, QCAL-IM, which upon proteolysis can be used as a universal ion mobility calibration standard. This single unique standard enables empirical calculation of peptide ion collision cross sections from the drift time on a T-Wave mobility instrument.

  16. A primary standard for low-g shock calibration by laser interferometry

    International Nuclear Information System (INIS)

    This paper presents a novel implementation of a primary standard for low-g shock acceleration calibration by laser interferometry based on rigid body collision at National Institute of Metrology, China. The mechanical structure of the standard device and working principles involved in the shock acceleration exciter, laser interferometers and virtual instruments are described. The novel combination of an electromagnetic exciter and a pneumatic exciter as the mechanical power supply of the standard device can deliver a wide range of shock acceleration levels. In addition to polyurethane rubber, two other types of material are investigated to ensure a wide selection of cushioning pads for shock pulse generation, with pulse shapes and data displayed. A heterodyne He–Ne laser interferometer is preferred for its precise and reliable measurement of shock acceleration while a homodyne one serves as a check standard. Some calibration results of a standard acceleration measuring chain are shown in company with the uncertainty evaluation budget. The expanded calibration uncertainty of shock sensitivity of the acceleration measuring chain is 0.8%, k = 2, with the peak acceleration range from 20 to 10 000 m s−2 and pulse duration from 0.5 to 10 ms. This primary shock standard can meet the traceability requirements of shock acceleration from various applications of industries from automobile to civil engineering and therefore is used for piloting the ongoing shock comparison of Technical Committee of Acoustics, Ultrasound and Vibration (TCAUV) of Asia Pacific Metrology Program (APMP), coded as APMP.AUV.V-P1. (paper)

  17. Dental Hygiene Faculty Calibration Using Two Accepted Standards for Calculus Detection: A Pilot Study.

    Science.gov (United States)

    Santiago, Lisa J; Freudenthal, Jacqueline J; Peterson, Teri; Bowen, Denise M

    2016-08-01

    Faculty calibration studies for calculus detection use two different standards for examiner evaluation, yet the only therapeutic modality that can be used for nonsurgical periodontal treatment is scaling/root debridement or planing. In this study, a pretest-posttest design was used to assess the feasibility of faculty calibration for calculus detection using two accepted standards: that established by the Central Regional Dental Testing Service, Inc. (CRDTS; readily detectible calculus) and the gold standard for scaling/root debridement (root roughness). Four clinical dental hygiene faculty members out of five possible participants at Halifax Community College agreed to participate. The participants explored calculus on the 16 assigned teeth (64 surfaces) of four patients. Calculus detection scores were calculated before and after training. Kappa averages using CRDTS criteria were 0.561 at pretest and 0.631 at posttest. Kappa scores using the scaling/root debridement or planing standard were 0.152 at pretest and 0.271 at posttest. The scores indicated improvement from moderate (Kappa=0.41-0.60) to substantial agreement (Kappa=0.61-0.80) following training using the CRDTS standard. Although this result differed qualitatively and Kappas were significantly different from 0, the differences for pre- to post-Kappas for patient-rater dyads using CRDTS were not statistically significant (p=0.778). There was no difference (p=0.913) in Kappa scores pre- to post-training using the scaling/root debridement standard. Despite the small number of participants in this study, the results indicated that training to improve interrater reliability to substantial agreement was feasible using the CRDTS standard but not using the gold standard. The difference may have been due to greater difficulty in attaining agreement regarding root roughness. Future studies should include multiple training sessions with patients using the same standard for scaling/root debridement used for

  18. Source geometry factors for HDR ¹⁹²Ir brachytherapy secondary standard well-type ionization chamber calibrations.

    Science.gov (United States)

    Shipley, D R; Sander, T; Nutbrown, R F

    2015-03-21

    Well-type ionization chambers are used for measuring the source strength of radioactive brachytherapy sources before clinical use. Initially, the well chambers are calibrated against a suitable national standard. For high dose rate (HDR) (192)Ir, this calibration is usually a two-step process. Firstly, the calibration source is traceably calibrated against an air kerma primary standard in terms of either reference air kerma rate or air kerma strength. The calibrated (192)Ir source is then used to calibrate the secondary standard well-type ionization chamber. Calibration laboratories are usually only equipped with one type of HDR (192)Ir source. If the clinical source type is different from that used for the calibration of the well chamber at the standards laboratory, a source geometry factor, k(sg), is required to correct the calibration coefficient for any change of the well chamber response due to geometric differences between the sources. In this work we present source geometry factors for six different HDR (192)Ir brachytherapy sources which have been determined using Monte Carlo techniques for a specific ionization chamber, the Standard Imaging HDR 1000 Plus well chamber with a type 70010 HDR iridium source holder. The calculated correction factors were normalized to the old and new type of calibration source used at the National Physical Laboratory. With the old Nucletron microSelectron-v1 (classic) HDR (192)Ir calibration source, ksg was found to be in the range 0.983 to 0.999 and with the new Isodose Control HDR (192)Ir Flexisource k(sg) was found to be in the range 0.987 to 1.004 with a relative uncertainty of 0.4% (k = 2). Source geometry factors for different combinations of calibration sources, clinical sources, well chambers and associated source holders, can be calculated with the formalism discussed in this paper.

  19. The development of an efficient mass balance approach for the purity assignment of organic calibration standards.

    Science.gov (United States)

    Davies, Stephen R; Alamgir, Mahiuddin; Chan, Benjamin K H; Dang, Thao; Jones, Kai; Krishnaswami, Maya; Luo, Yawen; Mitchell, Peter S R; Moawad, Michael; Swan, Hilton; Tarrant, Greg J

    2015-10-01

    The purity determination of organic calibration standards using the traditional mass balance approach is described. Demonstrated examples highlight the potential for bias in each measurement and the need to implement an approach that provides a cross-check for each result, affording fit for purpose purity values in a timely and cost-effective manner. Chromatographic techniques such as gas chromatography with flame ionisation detection (GC-FID) and high-performance liquid chromatography with UV detection (HPLC-UV), combined with mass and NMR spectroscopy, provide a detailed impurity profile allowing an efficient conversion of chromatographic peak areas into relative mass fractions, generally avoiding the need to calibrate each impurity present. For samples analysed by GC-FID, a conservative measurement uncertainty budget is described, including a component to cover potential variations in the response of each unidentified impurity. An alternative approach is also detailed in which extensive purification eliminates the detector response factor issue, facilitating the certification of a super-pure calibration standard which can be used to quantify the main component in less-pure candidate materials. This latter approach is particularly useful when applying HPLC analysis with UV detection. Key to the success of this approach is the application of both qualitative and quantitative (1)H NMR spectroscopy. PMID:26342310

  20. Calibrating and training of neutron based NSA techniques with less SNM standards

    Energy Technology Data Exchange (ETDEWEB)

    Geist, William H [Los Alamos National Laboratory; Swinhoe, Martyn T [Los Alamos National Laboratory; Bracken, David S [Los Alamos National Laboratory; Freeman, Corey R [Los Alamos National Laboratory; Newell, Matthew R [Los Alamos National Laboratory

    2010-01-01

    Accessing special nuclear material (SNM) standards for the calibration of and training on nondestructive assay (NDA) instruments has become increasingly difficult in light of enhanced safeguards and security regulations. Limited or nonexistent access to SNM has affected neutron based NDA techniques more than gamma ray techniques because the effects of multiplication require a range of masses to accurately measure the detector response. Neutron based NDA techniques can also be greatly affected by the matrix and impurity characteristics of the item. The safeguards community has been developing techniques for calibrating instrumentation and training personnel with dwindling numbers of SNM standards. Monte Carlo methods have become increasingly important for design and calibration of instrumentation. Monte Carlo techniques have the ability to accurately predict the detector response for passive techniques. The Monte Carlo results are usually benchmarked to neutron source measurements such as californium. For active techniques, the modeling becomes more difficult because of the interaction of the interrogation source with the detector and nuclear material; and the results cannot be simply benchmarked with neutron sources. A Monte Carlo calculated calibration curve for a training course in Indonesia of material test reactor (MTR) fuel elements assayed with an active well coincidence counter (AWCC) will be presented as an example. Performing training activities with reduced amounts of nuclear material makes it difficult to demonstrate how the multiplication and matrix properties of the item affects the detector response and limits the knowledge that can be obtained with hands-on training. A neutron pulse simulator (NPS) has been developed that can produce a pulse stream representative of a real pulse stream output from a detector measuring SNM. The NPS has been used by the International Atomic Energy Agency (IAEA) for detector testing and training applications at the

  1. Precise and accurate measurements of strong-field photoionisation and a transferrable laser intensity calibration standard

    CERN Document Server

    Wallace, W C; Khurmi, C; U., Satya Sainadh; Calvert, J E; Laban, D E; Pullen, M G; Bartschat, K; Grum-Grzhimailo, A N; Wells, D; Quiney, H M; Tong, X M; Litvinyuk, I V; Sang, R T; Kielpinski, D

    2016-01-01

    Ionization of atoms and molecules in strong laser fields is a fundamental process in many fields of research, especially in the emerging field of attosecond science. So far, demonstrably accurate data have only been acquired for atomic hydrogen (H), a species that is accessible to few investigators. Here we present measurements of the ionization yield for argon, krypton, and xenon with percentlevel accuracy, calibrated using H, in a laser regime widely used in attosecond science. We derive a transferrable calibration standard for laser peak intensity, accurate to 1.3%, that is based on a simple reference curve. In addition, our measurements provide a much-needed benchmark for testing models of ionisation in noble-gas atoms, such as the widely employed single-active electron approximation.

  2. Calibration of the Gaia RVS from ground-based observations of candidate standard stars

    OpenAIRE

    Chemin, L.; Soubiran, C.; Crifo, Françoise; Jasniewicz, Gérard; Katz, David; Hestroffer, Daniel; Udry, Stéphane

    2011-01-01

    International audience The Radial Velocity Spectrometer (RVS) on board of Gaia will perform a large spectroscopic survey to determine the radial velocities of some 1.5 × 10^8 stars. We present the status of ground-based observations of a sample of 1420 candidate standard stars designed to calibrate the RVS. Each candidate star has to be observed several times before Gaia launch (and at least once during the mission) to ensure that its radial velocity remains stable during the whole mission...

  3. Calibration of the Gaia Radial Velocity Spectrometer from ground-based observations of candidate standard stars

    OpenAIRE

    Chemin, L.; Soubiran, C.; Crifo, F.; Jasniewicz, G.; Katz, D.; Hestroffer, D.; Udry, S.

    2011-01-01

    The Radial Velocity Spectrometer (RVS) on board of Gaia will perform a large spectroscopic survey to determine the radial velocities of some 1.5x10^8 stars. We present the status of ground-based observations of a sample of 1420 candidate standard stars designed to calibrate the RVS. Each candidate star has to be observed several times before Gaia launch (and at least once during the mission) to ensure that its radial velocity remains stable during the whole mission. Observations are performed...

  4. Application of calibration standardization method to the analysis of diuretic pharmaceutical herbs

    International Nuclear Information System (INIS)

    Calibration standardization of X-ray fluorescence method was carried out for the determination of the anorganic contents in diuretic herbs as called Folia Betulae, Stylus Maydis, Flores Verbasci, Equisetum Arvense and Flos Helichrysi, growing in Turkey. These herbs are widely used in pharmacy and public health for kidney disease therapy. Herb samples were steeped in the water and mixed through the pure cellulose, then pelletized in the intermediate thickness. An annular source of 109Cd (3.7 MBq) was used for excitation of fluorescent K lines of elements lying between potassium and zirconium. Toxic elements in considerable amounts were not found. (author)

  5. Electromagnetic coupling into two standard calibration shields on the Sandia cable tester

    Energy Technology Data Exchange (ETDEWEB)

    Warne, Larry Kevin; Basilio, Lorena I.; Langston, William L.; Chen, Kenneth C.; Hudson, Howard Gerald; Morris, M. E.; Stronach, S.L.; Johnson, W. A.; Derr, W.

    2014-02-01

    This report presents analytic transmission line models for calculating the shielding effectiveness of two common calibration standard cables. The two cables have different canonical aperture types, which produce the same low frequency coupling but different responses at resonance. The dominant damping mechanism is produced by the current probe loads at the ends of the cables, which are characterized through adaptor measurements. The model predictions for the cables are compared with experimental measurements and good agreement between the results is demonstrated. This setup constitutes a nice repeatable geometry that nevertheless exhibits some of the challenges involved in modeling non-radio frequency geometries.

  6. X-ray and gamma-ray standards for detector calibration

    International Nuclear Information System (INIS)

    The IAEA established a Co-ordinated Research Programme (CRP) on the Measurements and Evaluation of X- and Gamma-Ray Standards for Detector Efficiency Calibration in 1986 with the aim of alleviating the generation of such discrepancies. Within the framework of this CRP, representatives of nine research groups from six Member States and one international organization performed a number of precise measurements and systematic in-depth evaluations of the required decay data. They have also contributed to the development of evaluation methodology and measurement techniques, and stimulated a number of such studies at laboratories not directly involved in the IAEA project. The results of the work of the CRP, which was finished in 1990, are presented in this report. Recommended values of half-lives and photon emission probabilities are given for a carefully selected set of radionuclides that are suitable for detector efficiency calibration (X-rays from 5 to 90 keV and gamma-rays from 30 to about 3000 keV). Detector efficiency calibration for higher gamma-ray energies (up to 14 MeV) is also considered. The evaluation procedures used to obtain the recommended values and their estimated uncertainties are reported, and a summary of the remaining discrepancies is given. Refs and tabs

  7. Interlaboratory comparison study of calibration standards for foraminiferal Mg/Ca thermometry

    Science.gov (United States)

    Greaves, M.; Caillon, N.; Rebaubier, H.; Bartoli, G.; Bohaty, S.; Cacho, I.; Clarke, L.; Cooper, M.; Daunt, C.; Delaney, M.; Demenocal, P.; Dutton, A.; Eggins, S.; Elderfield, H.; Garbe-Schoenberg, D.; Goddard, E.; Green, D.; Groeneveld, J.; Hastings, D.; Hathorne, E.; Kimoto, K.; Klinkhammer, G.; Labeyrie, L.; Lea, D. W.; Marchitto, T.; MartíNez-Botí, M. A.; Mortyn, P. G.; Ni, Y.; Nuernberg, D.; Paradis, G.; Pena, L.; Quinn, T.; Rosenthal, Y.; Russell, A.; Sagawa, T.; Sosdian, S.; Stott, L.; Tachikawa, K.; Tappa, E.; Thunell, R.; Wilson, P. A.

    2008-08-01

    An interlaboratory study of Mg/Ca and Sr/Ca ratios in three commercially available carbonate reference materials (BAM RS3, CMSI 1767, and ECRM 752-1) was performed with the participation of 25 laboratories that determine foraminiferal Mg/Ca ratios worldwide. These reference materials containing Mg/Ca in the range of foraminiferal calcite (0.8 mmol/mol to 6 mmol/mol) were circulated with a dissolution protocol for analysis. Participants were asked to make replicate dissolutions of the powdered samples and to analyze them using the instruments and calibration standards routinely used in their laboratories. Statistical analysis was performed in accordance with the International Standardization Organization standard 5725, which is based on the analysis of variance (ANOVA) technique. Repeatability (RSDr%), an indicator of intralaboratory precision, for Mg/Ca determinations in solutions after centrifuging increased with decreasing Mg/Ca, ranging from 0.78% at Mg/Ca = 5.56 mmol/mol to 1.15% at Mg/Ca = 0.79 mmol/mol. Reproducibility (RSDR%), an indicator of the interlaboratory method precision, for Mg/Ca determinations in centrifuged solutions was noticeably worse than repeatability, ranging from 4.5% at Mg/Ca = 5.56 mmol/mol to 8.7% at Mg/Ca = 0.79 mmol/mol. Results of this study show that interlaboratory variability is dominated by inconsistencies among instrument calibrations and highlight the need to improve interlaboratory compatibility. Additionally, the study confirmed the suitability of these solid standards as reference materials for foraminiferal Mg/Ca (and Sr/Ca) determinations, provided that appropriate procedures are adopted to minimize and to monitor possible contamination from silicate mineral phases.

  8. Airborne hygrometer calibration inter-comparison against a metrological water vapour standard

    Science.gov (United States)

    Smorgon, Denis; Boese, Norbert; Ebert, Volker

    2014-05-01

    Water vapour is the most important atmospheric greenhouse gas, which causes a major feedback to warming and other changes in the climate system. Knowledge of the distribution of water vapour and its climate induced changes is especially important in the upper troposphere and lower stratosphere (UT/LS) where vapour plays a critical role in atmospheric radiative balance, cirrus cloud formation, and photochemistry. But, our understanding of water in the UT/LS is limited by significant uncertainties in current UT/LS water measurements. One of the most comprehensive inter-comparison campaigns for airborne hygrometers, termed AQUAVIT (AV1) [1], took place in 2007 at the AIDA chamber at the Karlsruhe Institute of Technology (KIT) in Germany. AV1 was a well-defined, referred, blind inter-comparison of 22 airborne field instruments from 17 international research groups. One major metrological deficit of AV1, however, was, that no traceable reference instrument participated in the inter-comparison experiments and that the calibration procedures of the participating instruments were not monitored or interrogated. Consequently a follow-up inter-comparison was organized in April 2013, which for the first time also provides a traceable link to the international humidity scale. This AQUAVIT2 (AV2) campaign (details see: http://www.imk-aaf.kit.edu/aquavit/index.php/Main_Page) was again located at KIT/AIDA and organised by an international organizing committee including KIT, PTB, FZJ and others. Generally AV2 is divided in two parallel comparisons: 1) AV2-A uses the AIDA chamber for a simultaneous comparison of all instruments (incl. sampling and in-situ instruments) over a broad range of conditions characteristic for the UT/LS; 2) AV2-B, about which this paper is reporting, is a sequential comparison of selected hygrometers and (when possible) their reference calibration infrastructures by means of a chilled mirror hygrometer traced back to the primary National humidity standard

  9. A BAYESIAN METHOD FOR CALCULATING REAL-TIME QUANTITATIVE PCR CALIBRATION CURVES USING ABSOLUTE PLASMID DNA STANDARDS

    Science.gov (United States)

    In real-time quantitative PCR studies using absolute plasmid DNA standards, a calibration curve is developed to estimate an unknown DNA concentration. However, potential differences in the amplification performance of plasmid DNA compared to genomic DNA standards are often ignore...

  10. Calibration of the Gaia Radial Velocity Spectrometer from ground-based observations of candidate standard stars

    CERN Document Server

    Chemin, L; Crifo, F; Jasniewicz, G; Katz, D; Hestroffer, D; Udry, S

    2011-01-01

    The Radial Velocity Spectrometer (RVS) on board of Gaia will perform a large spectroscopic survey to determine the radial velocities of some 1.5x10^8 stars. We present the status of ground-based observations of a sample of 1420 candidate standard stars designed to calibrate the RVS. Each candidate star has to be observed several times before Gaia launch (and at least once during the mission) to ensure that its radial velocity remains stable during the whole mission. Observations are performed with the high-resolution spectrographs SOPHIE, NARVAL and CORALIE, completed with archival data of the ELODIE and HARPS instruments. The analysis shows that about 7% of the current catalogue exhibits variations larger than the adopted threshold of 300 m/s. Consequently, those stars should be rejected as reference targets, due to the expected accuracy of the Gaia RVS. Emphasis is also put here on our observations of bright asteroids to calibrate the ground-based velocities by a direct comparison with celestial mechanics. ...

  11. Final report on key comparison APMP.L-K4: Calibration of diameter standards

    Science.gov (United States)

    Chin, Jui-Hsi; Takatsuji, Toshiyuki; Horita, Masami; Hamakawa, Tsuyoshi; Chaudhary, K. P.; Tonmueanwai, Anusorn; Alfiyati, Nurul; Kruger, Oelof; Howick, Eleanor; Cox, Peter; Sawi, Muktar bin; Thu, Bui Quoc; Kim, Jong-Ahn; Yin, Wong Seung; Siew Leng, Tan; Zaher, S. Al

    2014-01-01

    A regional key comparison, APMP.L-K4, was held in 2008. To demonstrate the equivalence of routine calibration services offered by NMIs to clients, participants in this APMP.L-K4 comparison agreed to use the same apparatus and methods as routinely applied to client gauges. There are 14 laboratories from NMIs involved this key comparison, which included CMS/ITRI, NMIJ/AIST, NPL-I, NIMT, Puslit KIM-LIPI, NMISA, MSL, NMIA, NML-SIRIM, VMI, KRISS, SCL, NMC/A*STAR and NSCL. This report describes the measurement results of five diameter standards including two rings and three plugs. The calibrations of this key comparison were carried out by laboratories during the period from May 2008 to November 2010. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCL, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  12. Implementing a New Group of TPW Cells as National Standard: Impact on Calibration Services

    Science.gov (United States)

    Dobre, Miruna

    2008-06-01

    The definition of the kelvin is based on the triple-point temperature of highly pure water having the isotopic composition of ocean water (more specifically, the isotopic composition is equivalent to that of VSMOW). Belgian national metrology realizes the triple point of water (TPW) as the mean of temperatures measured in three sealed cells. In order to take into account the isotopic composition effect on TPW temperature, the ensemble of cells was replaced in 2006. Three new cells, with isotopic analysis of the contained water, were bought from different manufacturers. The new group of cells was compared to the old TPW national realization in order to quantify the effect of moving towards a new reference. Two different standard thermometers were used in all the cells to take 10 daily measurements on two different ice mantles. The measured resistances were corrected for hydrostatic head, self-heating, and isotopic composition (when available) before calculating the difference. A difference of about 87 μK was found between the old and the new national references. This difference is transferred to customers’ thermometers and cells through calibrations, and the change has to be documented in each new calibration certificate. An additional consequence of the new ensemble cell implementation is the significant reduction in the spread of deviations of individual cells from the mean temperature. The maximum difference between two cells of the ensemble is 96 μK for the old reference cells and 46 μK for the new reference cells corrected for isotopic composition effects.

  13. Binary Pseudo-random Grating Standard for Calibration of Surface Profilometers

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy; Yashchuk, Valeriy V.; McKinney, Wayne R.; Takacs, Peter Z.

    2008-01-16

    We suggest and describe the use of a binary pseudo-random (BPR) grating as a standard test surface for measurement of the modulation transfer function (MTF) of interferometric microscopes. Knowledge of the MTF of a microscope is absolutely necessary to convert the measured height distribution of a surface undergoing metrology into an accurate power spectral density (PSD) distribution. For an'ideal' microscope with an MTF function independent of spatial frequency out to the Nyquist frequency of the detector array with zero response at higher spatial frequencies, a BPR grating would produce a flat 1D PSD spectrum, independent of spatial frequency. For a'real' instrument, the MTF is found as the square root of the ratio of the PSD spectrum measured with the BPR grating to the'ideal,' spatial frequency independent, PSD spectrum. We present the results from a measurement of the MTF of MicromapTM-570 interferometric microscope demonstrating a high efficiency for the calibration method.

  14. Calibration and standards beamline 6.3.2 at the ALS

    Energy Technology Data Exchange (ETDEWEB)

    Underwood, J.H.; Gullikson, E.M.; Koike, M. [Ernest Orlando Lawrence Berkeley National Lab., CA (United States)] [and others

    1997-04-01

    More sophisticated optics for the x-ray, soft x-ray and far ultraviolet spectral regions being developed for synchrotron radiation research and many other applications, require accurate calibration and standards facilities for measuring reflectivity of mirrors and multilayer coatings, transmission of thin films, bandpass of multilayers, efficiency of gratings or detectors, etc. For this purpose beamline 6.3.2 was built at the ALS. Its energy coverage, versatility, simplicity and convenience also make it useful for a wide range of other experiments. The paper describes the components of this beamline, consisting of: a four jaw aperture; a horizontal focusing mirror; a monochromator; exit slit; vertical focusing mirror; mechanical and vacuum system; reflectometer; filter wheels; and data acquisition system.

  15. Automatable on-line generation of calibration curves and standard additions in solution-cathode glow discharge optical emission spectrometry

    International Nuclear Information System (INIS)

    Two methods are described that enable on-line generation of calibration standards and standard additions in solution-cathode glow discharge optical emission spectrometry (SCGD-OES). The first method employs a gradient high-performance liquid chromatography pump to perform on-line mixing and delivery of a stock standard, sample solution, and diluent to achieve a desired solution composition. The second method makes use of a simpler system of three peristaltic pumps to perform the same function of on-line solution mixing. Both methods can be computer-controlled and automated, and thereby enable both simple and standard-addition calibrations to be rapidly performed on-line. Performance of the on-line approaches is shown to be comparable to that of traditional methods of sample preparation, in terms of calibration curves, signal stability, accuracy, and limits of detection. Potential drawbacks to the on-line procedures include signal lag between changes in solution composition and pump-induced multiplicative noise. Though the new on-line methods were applied here to SCGD-OES to improve sample throughput, they are not limited in application to only SCGD-OES—any instrument that samples from flowing solution streams (flame atomic absorption spectrometry, ICP-OES, ICP-mass spectrometry, etc.) could benefit from them. - Highlights: • Describes rapid, on-line generation of calibration standards and standard additions • These methods enhance the ease of analysis and sample throughput with SCGD-OES. • On-line methods produce results comparable or superior to traditional calibration. • Possible alternative, null-point-based methods of calibration are described. • Methods are applicable to any system that samples from flowing liquid streams

  16. Non-matrix Matched Glass Disk Calibration Standards Improve XRF Micronutrient Analysis of Wheat Grain across Five Laboratories in India.

    Science.gov (United States)

    Guild, Georgia E; Stangoulis, James C R

    2016-01-01

    Within the HarvestPlus program there are many collaborators currently using X-Ray Fluorescence (XRF) spectroscopy to measure Fe and Zn in their target crops. In India, five HarvestPlus wheat collaborators have laboratories that conduct this analysis and their throughput has increased significantly. The benefits of using XRF are its ease of use, minimal sample preparation and high throughput analysis. The lack of commercially available calibration standards has led to a need for alternative calibration arrangements for many of the instruments. Consequently, the majority of instruments have either been installed with an electronic transfer of an original grain calibration set developed by a preferred lab, or a locally supplied calibration. Unfortunately, neither of these methods has been entirely successful. The electronic transfer is unable to account for small variations between the instruments, whereas the use of a locally provided calibration set is heavily reliant on the accuracy of the reference analysis method, which is particularly difficult to achieve when analyzing low levels of micronutrient. Consequently, we have developed a calibration method that uses non-matrix matched glass disks. Here we present the validation of this method and show this calibration approach can improve the reproducibility and accuracy of whole grain wheat analysis on 5 different XRF instruments across the HarvestPlus breeding program. PMID:27375644

  17. A Bayesian method for calculating real-time quantitative PCR calibration curves using absolute plasmid DNA standards

    Directory of Open Access Journals (Sweden)

    Haugland Richard A

    2008-02-01

    Full Text Available Abstract Background In real-time quantitative PCR studies using absolute plasmid DNA standards, a calibration curve is developed to estimate an unknown DNA concentration. However, potential differences in the amplification performance of plasmid DNA compared to genomic DNA standards are often ignored in calibration calculations and in some cases impossible to characterize. A flexible statistical method that can account for uncertainty between plasmid and genomic DNA targets, replicate testing, and experiment-to-experiment variability is needed to estimate calibration curve parameters such as intercept and slope. Here we report the use of a Bayesian approach to generate calibration curves for the enumeration of target DNA from genomic DNA samples using absolute plasmid DNA standards. Results Instead of the two traditional methods (classical and inverse, a Monte Carlo Markov Chain (MCMC estimation was used to generate single, master, and modified calibration curves. The mean and the percentiles of the posterior distribution were used as point and interval estimates of unknown parameters such as intercepts, slopes and DNA concentrations. The software WinBUGS was used to perform all simulations and to generate the posterior distributions of all the unknown parameters of interest. Conclusion The Bayesian approach defined in this study allowed for the estimation of DNA concentrations from environmental samples using absolute standard curves generated by real-time qPCR. The approach accounted for uncertainty from multiple sources such as experiment-to-experiment variation, variability between replicate measurements, as well as uncertainty introduced when employing calibration curves generated from absolute plasmid DNA standards.

  18. Standard practice for calibration of torque-measuring instruments for verifying the torque indication of torque testing machines

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2008-01-01

    1.1 This practice is to specify procedure for the calibration of elastic torque-measuring instruments. Note 1—Verification by deadweight and a lever arm is an acceptable method of verifying the torque indication of a torque testing machine. Tolerances for weights used are tabulated in Practice WK6364; methods for calibration of the weights are given in NIST Technical Note 577, Methods of Calibrating Weights for Piston Gages. 1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard. 1.3 This practice is intended for the calibration of static or quasi-static torque measuring instruments. The practice is not applicable for high speed torque calibrations or measurements. 1.4 This standard does not purport to address all of the safety concerns, if any,...

  19. Calibration standard of body tissue with magnetic nanocomposites for MRI and X-ray imaging

    Science.gov (United States)

    Rahn, Helene; Woodward, Robert; House, Michael; Engineer, Diana; Feindel, Kirk; Dutz, Silvio; Odenbach, Stefan; StPierre, Tim

    2016-05-01

    We present a first study of a long-term phantom for Magnetic Resonance Imaging (MRI) and X-ray imaging of biological tissues with magnetic nanocomposites (MNC) suitable for 3-dimensional and quantitative imaging of tissues after, e.g. magnetically assisted cancer treatments. We performed a cross-calibration of X-ray microcomputed tomography (XμCT) and MRI with a joint calibration standard for both imaging techniques. For this, we have designed a phantom for MRI and X-ray computed tomography which represents biological tissue enriched with MNC. The developed phantoms consist of an elastomer with different concentrations of multi-core MNC. The matrix material is a synthetic thermoplastic gel, PermaGel (PG). The developed phantoms have been analyzed with Nuclear Magnetic Resonance (NMR) Relaxometry (Bruker minispec mq 60) at 1.4 T to obtain R2 transverse relaxation rates, with SQUID (Superconducting QUantum Interference Device) magnetometry and Inductively Coupled Plasma Mass Spectrometry (ICP-MS) to verify the magnetite concentration, and with XμCT and 9.4 T MRI to visualize the phantoms 3-dimensionally and also to obtain T2 relaxation times. A specification of a sensitivity range is determined for standard imaging techniques X-ray computed tomography (XCT) and MRI as well as with NMR. These novel phantoms show a long-term stability over several months up to years. It was possible to suspend a particular MNC within the PG reaching a concentration range from 0 mg/ml to 6.914 mg/ml. The R2 relaxation rates from 1.4 T NMR-relaxometry show a clear connection (R2=0.994) with MNC concentrations between 0 mg/ml and 4.5 mg/ml. The MRI experiments have shown a linear correlation of R2 relaxation and MNC concentrations as well but in a range between MNC concentrations of 0 mg/ml and 1.435 mg/ml. It could be shown that XμCT displays best moderate and high MNC concentrations. The sensitivity range for this particular XμCT apparatus yields from 0.569 mg/ml to 6.914 mg/ml. The

  20. Calibration of a tertiary standard in N-ISO qualities for radioprotection and personal dosimetry

    International Nuclear Information System (INIS)

    Dosimetric calibration of radiation monitors and personal dosimeters in different radiological quantities are performed in order to obtain accurate measurements, for this reason the SSDL calculates the dosimetry calibration factor and its associated uncertainty, for each range of use. The calibration factor is performed using the known radiation field method and its uncertainty is calculated according to the ISO recommendations. The SSDL calculates the expanded uncertainty (Uc) with a coverage factor that provides a level of not less than 95 % of confidence. (authors).

  1. Development and validation of a portable gas phase standard generation and calibration system for volatile organic compounds

    Directory of Open Access Journals (Sweden)

    P. Veres

    2010-06-01

    Full Text Available We report on the development of an accurate, portable, dynamic calibration system for volatile organic compounds (VOCs. The Mobile Organic Carbon Calibration System (MOCCS combines the production of gas-phase VOC standards using permeation or diffusion sources with quantitative total organic carbon (TOC conversion on a palladium surface to CO2 in the presence of oxygen, and the subsequent CO2 measurement. MOCCS was validated using three different comparisons: (1 TOC of high accuracy methane standards compared well to expected concentrations (3% relative error, (2 a gas-phase benzene standard was generated using a permeation source and measured by TOC and gas chromatography mass spectrometry (GC-MS with excellent agreement (<4% relative difference, and (3 total carbon measurement of 4 known gas phase mixtures were performed and compared to a calculated carbon content to agreement within the stated uncertainties of the standards. Measurements from laboratory biomass burning experiments of formic acid by negative-ion proton-transfer chemical-ionization mass spectrometry (NI-PT-CIMS and formaldehyde by proton transfer reaction-mass spectrometry (PTR-MS, both calibrated using MOCCS, were compared to open path Fourier transform infrared spectroscopy (OP-FTIR to validate the MOCCS calibration and were found to compare well (R2 of 0.91 and 0.99, respectively.

  2. Development and validation of a portable gas phase standard generation and calibration system for volatile organic compounds

    Directory of Open Access Journals (Sweden)

    P. Veres

    2010-01-01

    Full Text Available We report on the development of an accurate, portable, dynamic calibration system for volatile organic compounds (VOCs. The Mobile Organic Carbon Calibration System (MOCCS combines the production of gas-phase VOC standards using permeation or diffusion sources with quantitative total organic carbon (TOC conversion on a palladium surface to CO2 in the presence of oxygen, and the subsequent CO2 measurement. MOCCS was validated using three different comparisons: (1 TOC of high accuracy methane standards compared well to expected concentrations (3% relative error, (2 a gas-phase benzene standard was generated using a permeation source and measured by TOC and gas chromatography mass spectrometry (GC-MS with excellent agreement (<4% relative difference, and (3 total carbon measurement of 4 known gas phase mixtures were performed and compared to a calculated carbon content to agreement within the stated uncertainties of the standards. Measurements from laboratory biomass burning experiments of formic acid by negative-ion proton-transfer chemical-ionization mass spectrometry (NI-PT-CIMS and formaldehyde by proton transfer reaction-mass spectrometry (PTR-MS, both calibrated using MOCCS, were compared to open path Fourier transform infrared spectroscopy (OP-FTIR to validate the MOCCS calibration and were found to compare well (R2 of 0.91 and 0.99 respectively.

  3. EPA Traceability Protocol for Assay and Certification of Gaseous Calibration Standards (EPA/600/R-12/531, May 2012)

    Science.gov (United States)

    In 1997, the U.S. Environmental Protection Agency (EPA) in Research Triangle Park, North Carolina, revised its 1993 version of its traceability protocol for the assay and certification of compressed gas and permeation-device calibration standards. The protocol allows producers of...

  4. The effects of ambient conditions on the calibration of air flow plate standards

    Directory of Open Access Journals (Sweden)

    Miao Qian

    2013-01-01

    Full Text Available The volume flow rate measured by air flow plate is influenced by the ambient conditions during the calibration. A series of numerical examples are conducted for the relationship and the outcomes demonstrated that the calibration is quite sensitive to the atmospheric pressure and the ambient temperature, but insensitive to relative humidity. The experiment model has been applied to calibration results with wide ranging ambient conditions. In conclusion, the results of this study demonstrate the benefits to calibration data of minimizing the effects of ambient conditions.

  5. Air index compensated interferometer as a prospective novel primary standard for baseline calibrations

    International Nuclear Information System (INIS)

    We describe the status of the development of an interferometer for absolute distance measurements with an intrinsic compensation of the refractive index of air, intended as a primary standard for the calibration of geodetic baselines. Two frequency-doubled Nd:YAG lasers are offset locked with a frequency difference of ≈20 GHz at the infrared 1064 nm fundamental wavelength. The resulting synthetic wavelengths of 15 mm for the infrared and 7.5 mm for the frequency-doubled green light are used as the scale for the measurements. Longer synthetic wavelengths are generated by acousto-optic frequency shifters. Based on the dispersion in air between green and infrared light the refractive index can be compensated. The attempt is demanding since uncertainties of the interferometric measurements for the optical wavelengths are scaled by a factor of nearly 300 000 in the refractive index compensated result. First comparisons up to 50 m length between this interferometer and a HeNe reference are presented. The deviations are smaller than ± 200 µm and dominated by a non-linearity from problems in the collimation of the measurement beams. In the linear parts the deviations are below ± 100 µm. (paper)

  6. Radioactivity in environmental samples: calibration standards measurement methods, quality assurance, and data analysis

    International Nuclear Information System (INIS)

    The numerous environmental radioactivity measurements made by and for the U.S. Environmental Protection Agency (U.S.EPA) include measurements on samples of water, urine, food, milk, and air filters. Calibration standards are listed which are available in the form of water solutions and soils for a wide range of radionuclides. Method validation procedures for U.S. EPA approval include protocol development and single-laboratory and multiple-laboratory evaluation for precision and accuracy. Inter-laboratory comparison studies are conducted for both cross-check and performance evaluation samples and involve 295 federal, state, and local laboratories. For water samples, 80% to 90% of the participating laboratories are within the control limits for most of the radionuclides measured; however, some problem areas exist, especially for radium-228 and strontium-89 and -90. For milk and food samples, more than 90% of the laboratories are within control limits for cobalt-60 and cesium-137 but some problems exist for the measurement of strontium-90, iodine-131, and potassium-40. For tritium, 91% of the laboratories are within the control limit for water samples and 87% are within the control limits for the urine samples. The laboratory performance for air filter samples shows some problems for gross beta, strontium-90 and cesium-137 measurements. (author)

  7. Calibration of Post-AGB Supergiants as Standard Extragalactic Candles for HST

    Science.gov (United States)

    Bond, Howard E.

    1998-01-01

    This report summarizes activities carried out with support from the NASA Ultraviolet, Visible, and Gravitational Astrophysics Research and Analysis Program. The aim of the program is to calibrate the absolute magnitudes of post-asymptotic-giant-branch (post-AGB or PAGB) stars, which we believe will be an excellent new "standard candle" for measuring extragalactic distances. The reason for this belief is that in old populations, the stars that are evolving through the PAGB region of the HR (Hertzsprung-Russell) diagram arise from only a single main-sequence turnoff mass. In addition, the theoretical PAGB evolutionary tracks show that they evolve through this region at constant luminosity; hence the PAGB stars should have an extremely narrow luminosity function. Moreover, as the PAGB stars evolve through spectral types F and A (en route from the AGB to hot stellar remnants and white dwarfs), they have the highest luminosities attained by old stars (both bolometrically and in the visual band). Finally, the PAGB stars of these spectral types are very easily identified, due to their large Balmer jumps, which are due to their very low surface gravities.

  8. Developing Calibration Weights and Standard-Error Estimates for a Survey of Drug-Related Emergency-Department Visits

    Directory of Open Access Journals (Sweden)

    Kott Phillip S.

    2014-09-01

    Full Text Available This article describes a two-step calibration-weighting scheme for a stratified simple random sample of hospital emergency departments. The first step adjusts for unit nonresponse. The second increases the statistical efficiency of most estimators of interest. Both use a measure of emergency-department size and other useful auxiliary variables contained in the sampling frame. Although many survey variables are roughly a linear function of the measure of size, response is better modeled as a function of the log of that measure. Consequently the log of size is a calibration variable in the nonresponse-adjustment step, while the measure of size itself is a calibration variable in the second calibration step. Nonlinear calibration procedures are employed in both steps. We show with 2010 DAWN data that estimating variances as if a one-step calibration weighting routine had been used when there were in fact two steps can, after appropriately adjusting the finite-population correct in some sense, produce standard-error estimates that tend to be slightly conservative.

  9. Development of transfer standard devices for ensuring the accurate calibration of ultrasonic physical therapy machines in clinical use

    International Nuclear Information System (INIS)

    Physical therapy ultrasound is widely applied to patients. However, many devices do not comply with the relevant standard stating that the actual power output shall be within ±20% of the device indication. Extreme cases have been reported: from delivering effectively no ultrasound or operating at maximum power at all powers indicated. This can potentially lead to patient injury as well as mistreatment. The present European (EC) project is an ongoing attempt to improve the quality of the treatment of patients being treated with ultrasonic physical-therapy. A Portable ultrasound Power Standard (PPS) is being developed and accurately calibrated. The PPS includes: Ultrasound transducers (including one exhibiting an unusual output) and a driver for the ultrasound transducers that has calibration and proficiency test functions. Also included with the PPS is a Cavitation Detector to determine the onset of cavitation occurring within the propagation medium. The PPS will be suitable for conducting in-the-field accreditation (proficiency testing and calibration). In order to be accredited it will be important to be able to show traceability of the calibration, the calibration process and qualification of testing staff. The clinical user will benefit from traceability because treatments will be performed more reliably

  10. Development of transfer standard devices for ensuring the accurate calibration of ultrasonic physical therapy machines in clinical use

    Energy Technology Data Exchange (ETDEWEB)

    Hekkenberg, R T [TNO Prevention and Health, Zernikedreef 9, 2333 CK Leiden (Netherlands); Richards, A [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Beissner, K [Physikalisch-Technische Bundesanstalt, PTB, Bundesallee 100, D-38116 Braunschweig (Germany); Zeqiri, B [National Physical Laboratory, NPL, Queens Road, Teddington, TW11 0LW (United Kingdom); Prout, G [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Cantrall, Ch [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Bezemer, R A [TNO Prevention and Health, Zernikedreef 9, 2333 CK Leiden (Netherlands); Koch, Ch [Physikalisch-Technische Bundesanstalt, PTB, Bundesallee 100, D-38116 Braunschweig, (Germany); Hodnett, M [National Physical Laboratory, NPL, Queens Road, Teddington, TW11 0LW (United Kingdom)

    2004-01-01

    Physical therapy ultrasound is widely applied to patients. However, many devices do not comply with the relevant standard stating that the actual power output shall be within {+-}20% of the device indication. Extreme cases have been reported: from delivering effectively no ultrasound or operating at maximum power at all powers indicated. This can potentially lead to patient injury as well as mistreatment. The present European (EC) project is an ongoing attempt to improve the quality of the treatment of patients being treated with ultrasonic physical-therapy. A Portable ultrasound Power Standard (PPS) is being developed and accurately calibrated. The PPS includes: Ultrasound transducers (including one exhibiting an unusual output) and a driver for the ultrasound transducers that has calibration and proficiency test functions. Also included with the PPS is a Cavitation Detector to determine the onset of cavitation occurring within the propagation medium. The PPS will be suitable for conducting in-the-field accreditation (proficiency testing and calibration). In order to be accredited it will be important to be able to show traceability of the calibration, the calibration process and qualification of testing staff. The clinical user will benefit from traceability because treatments will be performed more reliably.

  11. Standard Test Method for Calibration of Non-Concentrator Photovoltaic Secondary Reference Cells

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2010-01-01

    1.1 This test method covers calibration and characterization of secondary terrestrial photovoltaic reference cells to a desired reference spectral irradiance distribution. The recommended physical requirements for these reference cells are described in Specification E1040. Reference cells are principally used in the determination of the electrical performance of a photovoltaic device. 1.2 Secondary reference cells are calibrated indoors using simulated sunlight or outdoors in natural sunlight by reference to a primary reference cell previously calibrated to the same desired reference spectral irradiance distribution. 1.3 Secondary reference cells calibrated according to this test method will have the same radiometric traceability as the of the primary reference cell used for the calibration. Therefore, if the primary reference cell is traceable to the World Radiometric Reference (WRR, see Test Method E816), the resulting secondary reference cell will also be traceable to the WRR. 1.4 This test method appli...

  12. A new standard cylindrical graphite-walled ionization chamber for dosimetry in 60Co beams at calibration laboratories

    Science.gov (United States)

    Neves, Lucio P.; Perini, Ana P.; Caldas, Linda V. E.

    2014-11-01

    60Co sources are used mostly at dosimetry laboratories for calibration of ionization chambers utilized for radiotherapy dosimetry, mainly in those laboratories where there is no linear accelerator available. In this work, a new cylindrical ionization chamber was developed and characterized to be used as a reference dosimeter at the Calibration Laboratory of the IPEN. The characterization tests were performed according to the IEC 60731 standard, and all tests presented results within its recommended limits. Furthermore, the correction factors for the wall, stem, central collecting electrode, nonaxial uniformity and the mass-energy absorption coefficient were determined using the EGSnrc Monte Carlo code. The air kerma rate determined with this new dosimeter was compared to the one obtained with the IPEN standard, presenting a difference of 1.5%. Therefore, the new ionization chamber prototype developed and characterized in this work presents potential use as a primary standard dosimeter at radiation metrology laboratories.

  13. Flux Calibration of the BATC Survey and Its Application for Checking the Consistency of the Oke-Gunn Standards

    Institute of Scientific and Technical Information of China (English)

    2001-01-01

    Four Oke-Gunn (OG) standards, HD 19445, HD 84937, BD+26 2606 and BD+17 4708 are used as standard stars for flux calibration in the BATC project. They are also widely used in the visual wavelength region in many other photometric projects. Over the years we have observed on 58 good photometric nights, and the data obtained are used for flux calibration. Normally two or three OG standards are observed in every photometric night. The data are used for getting the atmo-spheric extinction coefficients and instrumental magnitude zero point. We also use these data to make inter-comparisons among the magnitudes of these standard stars.As a result, we found the magnitudes of HD 19445, HD 84937 and BD+17 4708 to agree well with those estimated in previous work to within 0.03 magnitude. How-ever, BD+26 2606 shows a larger deviation especially at short wavebands. Possible reasons are analyzed and the revised magnitudes are obtained for these standards.It is shown that the quality of flux calibration of the BATC fields is significantly improved by applying the new magnitudes.

  14. Spectral Irradiance Calibration in the Infrared. 11; Comparison of (alpha) Bootis and 1 Ceres with a Laboratory Standard

    Science.gov (United States)

    Witteborn, Fred C.; Cohen, Martin; Bregman, Jesse D.; Wooden, Diane H.; Heere, Karen; Shirley, Eric L.

    1999-01-01

    Infrared spectra of two celestial objects frequently used as flux standards are calibrated against an absolute laboratory flux standard at a spectral resolving power of 100 to 200. The spectrum of the KI.5 III star alpha Boo is measured from 3 to 30 microns, and that of the C-type asteroid 1 Ceres from 5 to 30 microns. While these "standard" spectra do not have the apparent precision of those based on calculated models, they do not require the assumptions involved in theoretical models of stars and asteroids. Specifically, they provide a model-independent means of calibrating celestial flux in the spectral range from 12 to 30 microns, where accurate absolute photometry is not available. The agreement found between the spectral shapes of alpha Boo and Ceres based on laboratory standards and those based on observed ratios to alpha CMa (Sirius) and alpha Lyr (Vega), flux-calibrated by theoretical modeling of these hot stars, strengthens our confidence in the applicability of the stellar models as primary irradiance standards.

  15. Spectral Irradiance Calibration in the Infrared 11: Comparison of (alpha) Boo and 1 Ceres with a Laboratory Standard

    Science.gov (United States)

    Witteborn, Fred C.; Cohen, Martin; Bregman, Jess D.; Wooden, Diane; Heere, Karen; Shirley, Eric L.

    1998-01-01

    Infrared spectra of two celestial objects frequently used as flux standards are calibrated against an absolute laboratory flux standard at a spectral resolving power of 100 to 200. The spectrum of the K1.5III star, alpha Boo, is measured from 3 microns to 30 microns and that of the C-type asteroid, 1 Ceres, from 5 microns to 30 microns. While these 'standard' spectra do not have the apparent precision of those based on calculated models, they do not require the assumptions involved in theoretical models of stars and asteroids. Specifically they provide a model-independent means of calibrating celestial flux in the spectral range from 12 microns to 30 microns where accurate absolute photometry is not available. The agreement found between the spectral shapes of alpha Boo and Ceres based on laboratory standards, and those based on observed ratios to alpha CMa (Sirius) and alpha Lyr (Vega), flux calibrated by theoretical modeling of these hot stars strengthens our confidence in the applicability of the stellar models as primary irradiance standards.

  16. Use of calibration standards and the correction for sample self-attenuation in gamma-ray nondestructive assay

    International Nuclear Information System (INIS)

    The efficient use of appropriate calibration standards and the correction for the attenuation of the gamma rays within an assay sample by the sample itself are two important and closely related subjects in gamma-ray nondestructive assay. Much research relating to those subjects has been done in the Nuclear Safeguards Research and Development program at the Los Alamos National Laboratory since 1970. This report brings together most of the significant results of that research. Also discussed are the nature of appropriate calibration standards and the necessary conditions on the composition, size, and shape of the samples to allow accurate assays. Procedures for determining the correction for the sample self-attenuation are described at length including both general principles and several specific useful cases. The most useful concept is that knowing the linear attenuation coefficient of the sample (which can usually be determined) and the size and shape of the sample and its position relative to the detector permits the computation of the correction factor for the self-attenuation. A major objective of the report is to explain how the procedures for determining the self-attenuation correction factor can be applied so that calibration standards can be entirely appropriate without being particularly similar, either physically or chemically, to the items to be assayed. This permits minimization of the number of standards required to assay items with a wide range of size, shape, and chemical composition. 17 references, 18 figures, 2 tables

  17. EASYTRAC Project: Work package 6.4 Reversal technique to calibrate gear and thread standards

    DEFF Research Database (Denmark)

    Carmignato, Simone; De Chiffre, Leonardo; Larsen, Erik;

    This report was produced as a part of the project EASYTRAC, an EU project under the programme Competitive and Sustainable Growth: Contract No. G6RD-CT-2000-00188, coordinated by UNIMETRIK S.A. (Spain). The project is concerned with low uncertainty calibrations on coordinate measuring machines (CM...... (PTB) - Germany and Tampere University of Technology (TUT) - Finland. The present report describes feasibility and experimental results of a reversal and substitute element technique application for thread calibration on CMMs....

  18. Linear model correction: A method for transferring a near-infrared multivariate calibration model without standard samples.

    Science.gov (United States)

    Liu, Yan; Cai, Wensheng; Shao, Xueguang

    2016-12-01

    Calibration transfer is essential for practical applications of near infrared (NIR) spectroscopy because the measurements of the spectra may be performed on different instruments and the difference between the instruments must be corrected. For most of calibration transfer methods, standard samples are necessary to construct the transfer model using the spectra of the samples measured on two instruments, named as master and slave instrument, respectively. In this work, a method named as linear model correction (LMC) is proposed for calibration transfer without standard samples. The method is based on the fact that, for the samples with similar physical and chemical properties, the spectra measured on different instruments are linearly correlated. The fact makes the coefficients of the linear models constructed by the spectra measured on different instruments are similar in profile. Therefore, by using the constrained optimization method, the coefficients of the master model can be transferred into that of the slave model with a few spectra measured on slave instrument. Two NIR datasets of corn and plant leaf samples measured with different instruments are used to test the performance of the method. The results show that, for both the datasets, the spectra can be correctly predicted using the transferred partial least squares (PLS) models. Because standard samples are not necessary in the method, it may be more useful in practical uses. PMID:27380302

  19. Test and Calibration of the Digital World-Wide Standardized Seismograph

    Science.gov (United States)

    Peterson, Jon; Hutt, Charles R.

    1982-01-01

    BACKGROUND During the past decade there has been steady progress in the modernization of the global seismograph network operated by the U.S. Geological Survey (USGS). The World-Wide Standardized Seismograph Network (WWSSN) has been augmented by new stations with advanced instrumentation, including the Seismic Research Observatories (SRO) and the modified High-Gain Long-Period (ASRO) stations. One goal in the modernization effort has been to improve signal resolution in the long-period band. A second goal has been to generate a global digital data base to support contemporary computer-based analysis and research. In 1976, a Panel on Seismograph Networks was established by the Committee on Seismology of the National Academy of Sciences to review progress in network seismology and recommend actions that would lead to an improved global data base for seismology. One recommendation in the Panel report (Engdahl, 1977) called for upgrading selected WWSSN stations by the installation of digital recorders. This was viewed as an economical way of expanding the digital network, which had proven itself to be a very promising new tool for earthquake and explosion research. Funds for the development and assembly of 15 digital recorders were provided to the USGS by the Defense Advanced Research Projects Agency and an ad hoc panel of scientists was convened by the Committee on Seismology to advise the USGS on the selection of stations to be upgraded and on data recording requirements. A total of 19 digital World-Wide Standardized Seismograph (DWWSS) systems will be operational when all are installed. The additional systems were made available through purchase by the USGS and other organizations; for example, the University of Bergen purchased and installed a DWWSS-type recorder and agreed to furnish the USGS with the data. A list of operational and planned DWWSS network stations is given in Table 1.1. As one might expect, the digital recorder turned out to be somewhat more

  20. Variations in performance of LCDs are still evident after DICOM gray-scale standard display calibration.

    LENUS (Irish Health Repository)

    Lowe, Joanna M

    2010-07-01

    Quality assurance in medical imaging is directly beneficial to image quality. Diagnostic images are frequently displayed on secondary-class displays that have minimal or no regular quality assurance programs, and treatment decisions are being made from these display types. The purpose of this study is to identify the impact of calibration on physical and psychophysical performance of liquid crystal displays (LCDs) and the extent of potential variance across various types of LCDs.

  1. Color standardization in whole slide imaging using a color calibration slide

    OpenAIRE

    Bautista, Pinky A.; Noriaki Hashimoto; Yukako Yagi

    2014-01-01

    Background: Color consistency in histology images is still an issue in digital pathology. Different imaging systems reproduced the colors of a histological slide differently. Materials and Methods: Color correction was implemented using the color information of the nine color patches of a color calibration slide. The inherent spectral colors of these patches along with their scanned colors were used to derive a color correction matrix whose coefficients were used to convert the pixels’ colors...

  2. Development of Multicolor Flow Cytometry Calibration Standards: Assignment of Equivalent Reference Fluorophores (ERF) Unit

    OpenAIRE

    Wang, Lili; Gaigalas, Adolfas K

    2011-01-01

    A procedure is described for assigning the number of equivalent reference fluorophores (ERF) values to microspheres labeled with a fluorophore designed to produce a fluorescence response in a given fluorescence channel of a multicolor flow cytometer. A fluorimeter was calibrated by a series of solutions of the reference fluorophores. The fluorimeter was used to obtain the microsphere fluorescence intensity, and a multicolor flow cytometer was used to obtain the microsphere concentration. The ...

  3. Preliminary Experiments to Develop a He-W Calibration Standard Using Laser Induced Breakdown Spectroscopy

    Science.gov (United States)

    Shaw, Guinevere; Andre, Nicolas; Bannister, Mark; Biewer, Theodore; Martin, Madhavi; Meyer, Fred; Wirth, Brian

    2015-11-01

    To address the needs of future fusion reactors, laser based diagnostic techniques for plasma-material interactions (PMI) are being developed at ORNL. Laser-induced breakdown spectroscopy (LIBS) is a technique for measuring elemental surface composition, and is a possible diagnostic for characterizing elemental concentrations in plasma-facing materials. The purpose of the LIBS system described here is to quantify helium (He) concentration in exposed tungsten (W) targets. To accurately quantify He concentration in situ a calibration stranded must be developed, including extensive calibration of the entire LIBS system. To accomplish this, two LIBS setups were explored: ex-situ LIBS and in-situ LIBS. Ex-situ LIBS experiments used W targets exposed to a He + ion beam to determine laser parameters and calibration settings for in-situ experiments. Results will be discussed. In-situ LIBS analysis will be assessed for W targets exposed to He plasma. Preliminary results will be discussed. Research sponsored by the Laboratory Directed Research and Development Program of Oak Ridge National Laboratory, managed by UT-Battelle, LLC, for the U. S. Department of Energy.

  4. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina

    2016-05-02

    This poster presents the development, implementation, and operation of the Broadband Outdoor Radiometer Calibrations (BORCAL) Longwave (LW) system at the Southern Great Plains Radiometric Calibration Facility for the calibration of pyrgeometers that provide traceability to the World Infrared Standard Group.

  5. Transition Radiation Tracker calibration, searches beyond the Standard Model and multiparticle correlations in ATLAS

    CERN Document Server

    Alonso, Alejandro; Torsten, Akesson

    This thesis contains two different aspects of my research work towards physics in proton-proton collisions in the ATLAS experiment at the LHC. The first part is focused on the understanding and developing of a calibration system to obtain the best possible charged particle reconstruction in the Transition Radiation Tracker. The method explained in this thesis is the current calibration technique used in the TRT and it is applied to all the data collected by ATLAS. Thanks to the method developed, the detector design resolution is achieved, and even improved in the central region of the TRT. In the second part, three different analyses are presented. Due to my interest in tracking and thanks to the new energy range available at the LHC, the first analysis is the study of multiparticle correlations at 900 GeV and 7 TeV. This analysis is performed with the first ATLAS data collected during 2010. Two different aspects are studied: the high order moments and an attempt to measure the normalized factorial moments ...

  6. (18)F primary standard at ENEA-INMRI by three absolute techniques and calibration of a well-type IG11 ionization chamber.

    Science.gov (United States)

    Capogni, Marco; Carconi, Pierluigi; De Felice, Pierino; Fazio, Aldo

    2016-03-01

    A new (18)F primary standardization carried out at ENEA-INMRI by three different absolute techniques, i.e. 4πγNaI(Tl)γ high-efficiency counting, TDCR and 4πβ(LS)-γ[NaI(Tl)] coincidence counting method, allowed the calibration of a fixed well-reentrant IG11 ionization chamber (IC), with an uncertainty lower than 1%, and to check the calibration factor of a portable well-type IC NPL-CRC model, previously calibrated. By the new standard the ENEA-INMRI was linked to the BIPM International Reference System (SIR) through the BIPM SIR Transfer Instrument (SIRTI).

  7. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    uncertainty was verified from independent measurements of the same sample by demonstrating statistical control of analytical results and the absence of bias. The proposed method takes into account uncertainties of the measurement, as well as of the amount of calibrant. It is applicable to all types......Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration...

  8. Oximetry system performance assessment with POM (acetal) phantoms incorporating hemoglobin calibration standards and customized saturation levels

    Science.gov (United States)

    Jang, Hyounguk; Singh, Karam; Wang, Hsing-Wen; Pfefer, T. J.; Chen, Yu

    2015-03-01

    Standardized approaches for performance assessment of biophotonic devices have the potential to facilitate system development and intercomparison, clinical trial standardization, recalibration, manufacturing quality control and quality assurance during clinical use. Evaluation of devices based on near-infrared spectroscopy (NIRS) for detection of hemoglobin (Hb) content and oxygenation have often involved tissue-simulating phantoms incorporating artificial dyes or flow systems. Towards the development of simple, effective techniques for objective, quantitative evaluation of basic NIRS system performance, we have developed and evaluated two test methods. These methods are based on cuvette inserts in solid turbid phantoms for measuring commercially-available Hb oximetry standards and custom-formulated oxy/deoxy-Hb solutions. Both approaches incorporate solid acetal, or polyoxymethylene (POM), as a tissue-simulating matrix material. First, inverse-adding-doubling (IAD) based on measurements with a spectrophotometer and an integrating sphere was used to measure POM optical properties and their stability over time. Second, two fiberopticprobe- based NIRS systems were used to measure concentration change of oxy- and deoxy-Hb in standard Hb solutions and customized Hb solutions by adding yeast. Differences in system performance were likely due to differences in light source outputs and fiberoptic probe design. Our preliminary results indicate that simple phantom-based approaches based on commercially available polymers and inclusions containing Hb standards, or controlled oxygenation levels may be useful for benchtop assessment of NIRS device quality for a variety of biophotonic devices.

  9. Development, improvement and calibration of neutronic reaction rate measurements: elaboration of a base of standard techniques; Developpement, amelioration et calibration des mesures de taux de reaction neutroniques: elaboration d`une base de techniques standards

    Energy Technology Data Exchange (ETDEWEB)

    Hudelot, J.P

    1998-06-19

    In order to improve and to validate the neutronic calculation schemes, perfecting integral measurements of neutronic parameters is necessary. This thesis focuses on the conception, the improvement and the development of neutronic reaction rates measurements, and aims at building a base of standard techniques. Two subjects are discussed. The first one deals with direct measurements by fission chambers. A short presentation of the different usual techniques is given. Then, those last ones are applied through the example of doubling time measurements on the EOLE facility during the MISTRAL 1 experimental programme. Two calibration devices of fission chambers are developed: a thermal column located in the central part of the MINERVE facility, and a calibration cell using a pulsed high flux neutron generator and based on the discrimination of the energy of the neutrons with a time-of-flight method. This second device will soon allow to measure the mass of fission chambers with a precision of about 1 %. Finally, the necessity of those calibrations will be shown through spectral indices measurements in core MISTRAL 1 (UO{sub 2}) and MISTRAL 2 (MOX) of the EOLE facility. In each case, the associated calculation schemes, performed using the Monte Carlo MCNP code with the ENDF-BV library, will be validated. Concerning the second one, the goal is to develop a method for measuring the modified conversion ratio of {sup 238}U (defined as the ratio of {sup 238}U capture rate to total fission rate) by gamma-ray spectrometry of fuel rods. Within the framework of the MISTRAL 1 and MISTRAL 2 programmes, the measurement device, the experimental results and the spectrometer calibration are described. Furthermore, the MCNP calculations of neutron self-shielding and gamma self-absorption are validated. It is finally shown that measurement uncertainties are better than 1 %. The extension of this technique to future modified conversion ratio measurements for {sup 242}Pu (on MOX rods) and

  10. Free-field reciprocity calibration of laboratory standard (LS) microphones using a time selective technique

    DEFF Research Database (Denmark)

    Rasmussen, Knud; Barrera Figueroa, Salvador

    2006-01-01

    sensitivities. Next an inverse FFT is applied and a time window around the main signal is used to eliminate cross talk and reflections. Finally, the signal is transformed back to the frequency domain and the free field sensitivities calculated. The standard procedure at DPLA involves measurements at four...

  11. Contributed Review: Absolute spectral radiance calibration of fiber-optic shock-temperature pyrometers using a coiled-coil irradiance standard lamp

    Science.gov (United States)

    Fat'yanov, O. V.; Asimow, P. D.

    2015-10-01

    We describe an accurate and precise calibration procedure for multichannel optical pyrometers such as the 6-channel, 3-ns temporal resolution instrument used in the Caltech experimental geophysics laboratory. We begin with a review of calibration sources for shock temperatures in the 3000-30 000 K range. High-power, coiled tungsten halogen standards of spectral irradiance appear to be the only practical alternative to NIST-traceable tungsten ribbon lamps, which are no longer available with large enough calibrated area. However, non-uniform radiance complicates the use of such coiled lamps for reliable and reproducible calibration of pyrometers that employ imaging or relay optics. Careful analysis of documented methods of shock pyrometer calibration to coiled irradiance standard lamps shows that only one technique, not directly applicable in our case, is free of major radiometric errors. We provide a detailed description of the modified Caltech pyrometer instrument and a procedure for its absolute spectral radiance calibration, accurate to ±5%. We employ a designated central area of a 0.7× demagnified image of a coiled-coil tungsten halogen lamp filament, cross-calibrated against a NIST-traceable tungsten ribbon lamp. We give the results of the cross-calibration along with descriptions of the optical arrangement, data acquisition, and processing. We describe a procedure to characterize the difference between the static and dynamic response of amplified photodetectors, allowing time-dependent photodiode correction factors for spectral radiance histories from shock experiments. We validate correct operation of the modified Caltech pyrometer with actual shock temperature experiments on single-crystal NaCl and MgO and obtain very good agreement with the literature data for these substances. We conclude with a summary of the most essential requirements for error-free calibration of a fiber-optic shock-temperature pyrometer using a high-power coiled tungsten halogen

  12. Contributed Review: Absolute spectral radiance calibration of fiber-optic shock-temperature pyrometers using a coiled-coil irradiance standard lamp

    Energy Technology Data Exchange (ETDEWEB)

    Fat’yanov, O. V., E-mail: fatyan1@gps.caltech.edu; Asimow, P. D., E-mail: asimow@gps.caltech.edu [Division of Geological and Planetary Sciences 252-21, California Institute of Technology, Pasadena, California 91125 (United States)

    2015-10-15

    We describe an accurate and precise calibration procedure for multichannel optical pyrometers such as the 6-channel, 3-ns temporal resolution instrument used in the Caltech experimental geophysics laboratory. We begin with a review of calibration sources for shock temperatures in the 3000-30 000 K range. High-power, coiled tungsten halogen standards of spectral irradiance appear to be the only practical alternative to NIST-traceable tungsten ribbon lamps, which are no longer available with large enough calibrated area. However, non-uniform radiance complicates the use of such coiled lamps for reliable and reproducible calibration of pyrometers that employ imaging or relay optics. Careful analysis of documented methods of shock pyrometer calibration to coiled irradiance standard lamps shows that only one technique, not directly applicable in our case, is free of major radiometric errors. We provide a detailed description of the modified Caltech pyrometer instrument and a procedure for its absolute spectral radiance calibration, accurate to ±5%. We employ a designated central area of a 0.7× demagnified image of a coiled-coil tungsten halogen lamp filament, cross-calibrated against a NIST-traceable tungsten ribbon lamp. We give the results of the cross-calibration along with descriptions of the optical arrangement, data acquisition, and processing. We describe a procedure to characterize the difference between the static and dynamic response of amplified photodetectors, allowing time-dependent photodiode correction factors for spectral radiance histories from shock experiments. We validate correct operation of the modified Caltech pyrometer with actual shock temperature experiments on single-crystal NaCl and MgO and obtain very good agreement with the literature data for these substances. We conclude with a summary of the most essential requirements for error-free calibration of a fiber-optic shock-temperature pyrometer using a high-power coiled tungsten halogen

  13. Contributed Review: Absolute spectral radiance calibration of fiber-optic shock-temperature pyrometers using a coiled-coil irradiance standard lamp

    International Nuclear Information System (INIS)

    We describe an accurate and precise calibration procedure for multichannel optical pyrometers such as the 6-channel, 3-ns temporal resolution instrument used in the Caltech experimental geophysics laboratory. We begin with a review of calibration sources for shock temperatures in the 3000-30 000 K range. High-power, coiled tungsten halogen standards of spectral irradiance appear to be the only practical alternative to NIST-traceable tungsten ribbon lamps, which are no longer available with large enough calibrated area. However, non-uniform radiance complicates the use of such coiled lamps for reliable and reproducible calibration of pyrometers that employ imaging or relay optics. Careful analysis of documented methods of shock pyrometer calibration to coiled irradiance standard lamps shows that only one technique, not directly applicable in our case, is free of major radiometric errors. We provide a detailed description of the modified Caltech pyrometer instrument and a procedure for its absolute spectral radiance calibration, accurate to ±5%. We employ a designated central area of a 0.7× demagnified image of a coiled-coil tungsten halogen lamp filament, cross-calibrated against a NIST-traceable tungsten ribbon lamp. We give the results of the cross-calibration along with descriptions of the optical arrangement, data acquisition, and processing. We describe a procedure to characterize the difference between the static and dynamic response of amplified photodetectors, allowing time-dependent photodiode correction factors for spectral radiance histories from shock experiments. We validate correct operation of the modified Caltech pyrometer with actual shock temperature experiments on single-crystal NaCl and MgO and obtain very good agreement with the literature data for these substances. We conclude with a summary of the most essential requirements for error-free calibration of a fiber-optic shock-temperature pyrometer using a high-power coiled tungsten halogen

  14. The ACS survey of globular clusters. XIII. Photometric calibration in comparison with Stetson standards

    CERN Document Server

    Hempel, Maren; Anderson, Jay; Aparicio, Antonio; Bedin, Luigi R; Chaboyer, Brian; Majewski, Steven R; Marín-Franch, Antonio; Milone, Antonino; Paust, Nathaniel E Q; Piotto, Giampaolo; Reid, I Neill; Rosenberg, Alfred; Siegel, Michael

    2013-01-01

    In this study we compare the photometric data of 34 Milky Way globular clusters, observed within the ACS Treasury Program (PI: Ata Sarajedini) with the corresponding ground-based data, provided by the Photometric Standard Field Catalogs of Stetson (2000, 2005). We focus on the transformation between the HST/ACS F606W to V-band and F814W to I-band only. The goal is to assess the validity of the filter transformation equations by Sirianni et al.(2005) with respect to their dependence on metallicity, Horizontal Branch morphology, mass and integrated (V-I) colour of the various globular clusters. Such a dependence is expected due to the fact that the transformation equations are based on the observations of only one globular cluster, i.e., NGC 2419. Surprisingly, the correlation between offset and metallicity is found to be weak, with a low level significance. The correlation between offset and Horizontal Branch structure, as well as total cluster mass is still weaker. Based on the available data we do not find t...

  15. CRM Assessment: Determining the Generalization of Rater Calibration Training. Summary of Research Report: Gold Standards Training

    Science.gov (United States)

    Baker, David P.

    2002-01-01

    The extent to which pilot instructors are trained to assess crew resource management (CRM) skills accurately during Line-Oriented Flight Training (LOFT) and Line Operational Evaluation (LOE) scenarios is critical. Pilot instructors must make accurate performance ratings to ensure that proper feedback is provided to flight crews and appropriate decisions are made regarding certification to fly the line. Furthermore, the Federal Aviation Administration's (FAA) Advanced Qualification Program (AQP) requires that instructors be trained explicitly to evaluate both technical and CRM performance (i.e., rater training) and also requires that proficiency and standardization of instructors be verified periodically. To address the critical need for effective pilot instructor training, the American Institutes for Research (AIR) reviewed the relevant research on rater training and, based on "best practices" from this research, developed a new strategy for training pilot instructors to assess crew performance. In addition, we explored new statistical techniques for assessing the effectiveness of pilot instructor training. The results of our research are briefly summarized below. This summary is followed by abstracts of articles and book chapters published under this grant.

  16. A comparative study between different alternatives to prepare gaseous standards for calibrating UV-Ion Mobility Spectrometers.

    Science.gov (United States)

    Criado-García, Laura; Garrido-Delgado, Rocío; Arce, Lourdes; Valcárcel, Miguel

    2013-07-15

    An UV-Ion Mobility Spectrometer is a simple, rapid, inexpensive instrument widely used in environmental analysis among other fields. The advantageous features of its underlying technology can be of great help towards developing reliable, economical methods for determining gaseous compounds from gaseous, liquid and solid samples. Developing an effective method using UV-Ion Mobility Spectrometry (UV-IMS) to determine volatile analytes entails using appropriate gaseous standards for calibrating the spectrometer. In this work, two home-made sample introduction systems (SISs) and a commercial gas generator were used to obtain such gaseous standards. The first home-made SIS used was a static head-space to measure compounds present in liquid samples and the other home-made system was an exponential dilution set-up to measure compounds present in gaseous samples. Gaseous compounds generated by each method were determined on-line by UV-IMS. Target analytes chosen for this comparative study were ethanol, acetone, benzene, toluene, ethylbenzene and xylene isomers. The different alternatives were acceptable in terms of sensitivity, precision and selectivity.

  17. THE ACS SURVEY OF GLOBULAR CLUSTERS. XIII. PHOTOMETRIC CALIBRATION IN COMPARISON WITH STETSON STANDARDS

    Energy Technology Data Exchange (ETDEWEB)

    Hempel, Maren [Instituto de Astrofísica, Facultad de Fisica, Pontificia Universidad Católica de Chile, Vicuña Mackenna 4860, 7820436 Macul, Santiago (Chile); Sarajedini, Ata [Department of Astronomy, University of Florida, 211 Bryant Space Science Center, Gainesville, FL 32611 (United States); Anderson, Jay; Reid, I. Neill, E-mail: mhempel@astro.puc.cl, E-mail: ata@astro.ufl.edu, E-mail: jayander@stsci.edu, E-mail: inr@stsci.edu [Space Telescope Science Institute, 3700 San Martin Drive, Baltimore, MD 21218 (United States); and others

    2014-03-01

    In this study we compare the photometric data of 34 Milky Way globular clusters, observed within the Advanced Camera for Surveys (ACS) Treasury Program (PI: A. Sarajedini) with the corresponding ground-based data, provided by the Photometric Standard Field Catalogs of Stetson. We focus on the transformation between the Hubble Space Telescope/ACS F606W to V-band and F814W to I-band only. The goal is to assess the validity of the filter transformation equations by Sirianni et al. with respect to their dependence on metallicity, horizontal branch morphology, mass, and integrated (V – I) color of the various globular clusters. The transformation equations as recommended by Sirianni et al. are based on synthetic photometry, were mostly tested on NGC 2419, and may introduce additional uncertainties when applied to different stellar populations. Such a dependence is expected due to the fact that the transformation equations are based on the observations of only one globular cluster, i.e., NGC 2419. Surprisingly, the correlation between offset and metallicity is found to be weak, with a low level significance. The correlation between offset and horizontal branch structure, as well as total cluster mass is still weaker. Based on the available data we do not find the photometric offset to be linked to multiple stellar populations, e.g., as found in NGC 0288, NGC 1851, and NGC 5139. The results of this study show that there are small systematic offsets between the transformed ACS- and observed ground-based photometry, and that these are only weakly correlated, if at all, with various cluster parameters and their underlying stellar populations. As a result, investigators wishing to transform globular cluster photometry from the Sirianni et al. ground-based V, I system onto the Stetson system simply need to add –0.040 (±0.012) to the V magnitudes and –0.047 (±0.011) to the I magnitudes. This in turn means that the transformed ACS V – I colors match the ground

  18. Radiometric calibration of optical microscopy and microspectroscopy apparata over a broad spectral range using a special thin-film luminescence standard

    Directory of Open Access Journals (Sweden)

    J. Valenta

    2015-04-01

    Full Text Available Application capabilities of optical microscopes and microspectroscopes can be considerably enhanced by a proper calibration of their spectral sensitivity. We propose and demonstrate a method of relative and absolute calibration of a microspectroscope over an extraordinary broad spectral range covered by two (parallel detection branches in visible and near-infrared spectral regions. The key point of the absolute calibration of a relative spectral sensitivity is application of the standard sample formed by a thin layer of Si nanocrystals with stable and efficient photoluminescence. The spectral PL quantum yield and the PL spatial distribution of the standard sample must be characterized by separate experiments. The absolutely calibrated microspectroscope enables to characterize spectral photon emittance of a studied object or even its luminescence quantum yield (QY if additional knowledge about spatial distribution of emission and about excitance is available. Capabilities of the calibrated microspectroscope are demonstrated by measuring external QY of electroluminescence from a standard poly-Si solar-cell and of photoluminescence of Er-doped Si nanocrystals.

  19. Weight-based synthesized standards preparation for correction-free calibration in X-ray fluorescence determination of tungsten in high-speed steel

    International Nuclear Information System (INIS)

    This paper suggests a correction-free calibration method in wavelength dispersive X-ray fluorescence analysis in order to determine tungsten as a major alloyed element in high-speed steels accurately. Matrix effects on fluorescent X-ray intensity of tungsten Lα line were minimized by borate fusion, and the total amount of tungsten in the glassy matrix could be quantified. Glass bead specimens were prepared with 10 to 12 mg of the steel sample and 4.0 g of lithium tetraborate as a flux agent. Without untraceable X-ray intensity correction, a linear calibration curve was obtained by measuring synthesized calibration standards prepared by using standard solutions. As compared with fundamental parameter calculations, the present method gave more accurate results of tungsten in certified reference materials of high-speed steel. (author)

  20. The standardization methods of radioactive sources (125I, 131I, 99mTc, and 18F) for calibrating nuclear medicine equipment in Indonesia

    Science.gov (United States)

    Wurdiyanto, G.; Candra, H.

    2016-03-01

    The standardization of radioactive sources (125I, 131I, 99mTc and 18F) to calibrate the nuclear medicine equipment had been carried out in PTKMR-BATAN. This is necessary because the radioactive sources used in the field of nuclear medicine has a very short half-life in other that to obtain a quality measurement results require special treatment. Besides that, the use of nuclear medicine techniques in Indonesia develop rapidly. All the radioactive sources were prepared by gravimetric methods. Standardization of 125I has been carried out by photon- photon coincidence methods, while the others have been carried out by gamma spectrometry methods. The standar sources are used to calibrate a Capintec CRC-7BT radionuclide calibrator. The results shows that calibration factor for Capintec CRC-7BT dose calibrator is 1,03; 1,02; 1,06; and 1,04 for 125I, 131I, 99mTc and 18F respectively, by about 5 to 6% of the expanded uncertainties.

  1. Using CEOS reference standard test sites to track the calibration stability of NOAA-19 AVHRR reflective solar channels

    Science.gov (United States)

    Wu, Aisheng; Angal, Amit; Xiong, Jack; Cao, Changyong

    2010-10-01

    In recent years, there is an increasing interest to establish a global integrated network of calibration sites for the purpose of tracking sensor performance, conducting cross-sensor comparison and assessing data quality and consistency. Based on such a need, the Committee on Earth Observation Satellites (CEOS) proposed eight instrumented sites for which surface measurements can be acquired through field campaigns and five pseudo-invariant desert sites typically consisting of sand dunes. In this study, we select one site from each category to study the calibration stability of reflective solar channels of NOAA- 19 Advanced Very High Resolution Radiometer (AVHRR) (launched on February 6, 2009). Since AVHRR does not have an onboard calibrator for the reflective solar channels and vicarious calibration often needs long-term observations to derive reliable trends, this study will provide an early assessment of sensor on-orbit calibration performance and establish a preliminary trend to examine its calibration consistency with other sensors. The Antarctic Dome C site is selected primarily to monitor the on-orbit calibration performance whereas Libya 4 test site is used to evaluate the cross-calibration consistency of AVHRR with other sensors. A site-specific Bi-directional Reflectance Distribution Function (BRDF) model developed based on observations made by Moderate Resolution Imaging Spectroradiometer (MODIS) is used to normalize AVHRR observed Top-of-Atmosphere (TOA) reflectances. Impact due to calibration applied to NOAA-19 AVHRR L1B is assessed separately using a constant detector response. Results show that for NOAA-19 AVHRR solar channels 1 and 2, variations in reflectance during the first year after launch are still around 6% and more than 10%, respectively, either due to sensor change or improper adjustment of calibration coefficients. While two sites provide consistent trends for the visible channel, the Dome C site is more suitable for the near

  2. A novel strategy for preparing calibration standards for the analysis of plant materials by laser-induced breakdown spectroscopy: A case study with pellets of sugar cane leaves

    Energy Technology Data Exchange (ETDEWEB)

    Silva Gomes, Marcos da [Universidade de São Paulo, Centro de Energia Nuclear na Agricultura (CENA), Laboratório de Química Analítica, Caixa Postal 96, CEP 13416-000, Piracicaba, SP (Brazil); Universidade Federal de São Carlos, Departamento de Química, Rodovia Washington Luiz km 235, São Carlos, SP (Brazil); Gustinelli Arantes de Carvalho, Gabriel [Universidade de São Paulo, Centro de Energia Nuclear na Agricultura (CENA), Laboratório de Química Analítica, Caixa Postal 96, CEP 13416-000, Piracicaba, SP (Brazil); Santos, Dário [Universidade Federal de São Paulo, Departamento de Ciências Exatas e da Terra, Rua Professor Arthur Riedel 275, Diadema, SP (Brazil); Krug, Francisco José, E-mail: fjkrug@cena.usp.br [Universidade de São Paulo, Centro de Energia Nuclear na Agricultura (CENA), Laboratório de Química Analítica, Caixa Postal 96, CEP 13416-000, Piracicaba, SP (Brazil)

    2013-08-01

    Calibration is still a challenging task when dealing with the direct analysis of solids. This is particularly true for laser-induced breakdown spectroscopy (LIBS), and laser ablation inductively coupled plasma optical emission spectrometry/mass spectrometry, when the calibrations are matrix-dependent and/or appropriate certified reference materials are generally not available. Looking at the analysis of plant materials in the form of pressed pellets by LIBS, a new method to overcome and/or minimize this difficulty is proposed by keeping the matrix constant in order to produce matrix-matched calibration pellets. To achieve this goal and to test this novel approach, ground sugar cane leaves were chosen and submitted to acid extractions for obtaining the corresponding blank or a material containing very low concentrations of the analytes. The resulting dried solid material was used either as a blank or a low concentration standard, and also homogeneously mixed with the original plant material at appropriate ratios as well. The corresponding pellets were used as calibration standards and ablated at 30 different sites by applying 25 laser pulses per site with a Q-switched Nd:YAG at 1064 nm. The plasma emission collected by lenses was directed through an optical fiber towards a spectrometer equipped with Echelle optics and intensified charge-coupled device. Delay time and integration time gate were fixed at 2.0 and 5.0 μs, respectively. This calibration strategy was tested for the determination of Ca, Mg, K, P, Cu, Mn, and Zn by LIBS in pellets of leaves from 17 varieties of sugar cane and good correlations were obtained with inductively coupled plasma optical emission spectrometry results in the corresponding acid digests. The proposed approach was also useful to estimate the limits of detection based on measurements of blanks, as recommended by IUPAC, or with the aid of a low concentration standard. - Highlights: • Blanks and/or low concentration standards of plant

  3. NBS (National Bureau of Standards) measurement services: calibration of beta-particle-emitting ophthalmic applicators. Final report

    International Nuclear Information System (INIS)

    The service provides calibrations for (90)Sr + (90)Y beta-particle ophthalmic applicators. The calibration determines the average surface absorbed-dose rate to water over the active area of the applicator. The technique used is to measure current per unit mass of air at the active surface with an extrapolation ionization chamber, and to convert this into absorbed-dose rate with Bragg-Gray cavity ionization theory. The extrapolation chamber measurements are made in three parts. Data-book measurements and a calibration report are given for one particular applicator

  4. Standard Test Method for Calibration of Primary Non-Concentrator Terrestrial Photovoltaic Reference Cells Using a Tabular Spectrum

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2010-01-01

    1.1 This test method is intended to be used for calibration and characterization of primary terrestrial photovoltaic reference cells to a desired reference spectral irradiance distribution, such as Tables G173. The recommended physical requirements for these reference cells are described in Specification E1040. Reference cells are principally used in the determination of the electrical performance of photovoltaic devices. 1.2 Primary photovoltaic reference cells are calibrated in natural sunlight using the relative spectral response of the cell, the relative spectral distribution of the sunlight, and a tabulated reference spectral irradiance distribution. 1.3 This test method requires the use of a pyrheliometer that is calibrated according to Test Method E816, which requires the use of a pyrheliometer that is traceable to the World Radiometric Reference (WRR). Therefore, reference cells calibrated according to this test method are traceable to the WRR. 1.4 This test method is a technique that may be used ...

  5. A novel strategy for preparing calibration standards for the analysis of plant materials by laser-induced breakdown spectroscopy: A case study with pellets of sugar cane leaves

    Science.gov (United States)

    da Silva Gomes, Marcos; de Carvalho, Gabriel Gustinelli Arantes; Santos, Dário, Junior; Krug, Francisco José

    2013-08-01

    Calibration is still a challenging task when dealing with the direct analysis of solids. This is particularly true for laser-induced breakdown spectroscopy (LIBS), and laser ablation inductively coupled plasma optical emission spectrometry/mass spectrometry, when the calibrations are matrix-dependent and/or appropriate certified reference materials are generally not available. Looking at the analysis of plant materials in the form of pressed pellets by LIBS, a new method to overcome and/or minimize this difficulty is proposed by keeping the matrix constant in order to produce matrix-matched calibration pellets. To achieve this goal and to test this novel approach, ground sugar cane leaves were chosen and submitted to acid extractions for obtaining the corresponding blank or a material containing very low concentrations of the analytes. The resulting dried solid material was used either as a blank or a low concentration standard, and also homogeneously mixed with the original plant material at appropriate ratios as well. The corresponding pellets were used as calibration standards and ablated at 30 different sites by applying 25 laser pulses per site with a Q-switched Nd:YAG at 1064 nm. The plasma emission collected by lenses was directed through an optical fiber towards a spectrometer equipped with Echelle optics and intensified charge-coupled device. Delay time and integration time gate were fixed at 2.0 and 5.0 μs, respectively. This calibration strategy was tested for the determination of Ca, Mg, K, P, Cu, Mn, and Zn by LIBS in pellets of leaves from 17 varieties of sugar cane and good correlations were obtained with inductively coupled plasma optical emission spectrometry results in the corresponding acid digests. The proposed approach was also useful to estimate the limits of detection based on measurements of blanks, as recommended by IUPAC, or with the aid of a low concentration standard.

  6. Effect of storage regime on the stability of DNA used as a calibration standard for real-time polymerase chain reaction.

    Science.gov (United States)

    Podivinsky, Ellen; Love, John L; van der Colff, Loraine; Samuel, Laly

    2009-11-01

    This article looks at storage factors influencing the stability of potential DNA calibration standards for use in quantitative polymerase chain reaction (PCR). Target sequences from the bacteria Campylobacter jejuni were cloned into a plasmid vector. Samples of these potential calibration standards were stored at +4, -20, and -80 degrees C as aqueous and lyophilized samples and were prepared as both single-use aliquots and multiple-use preparations. Results showed that the samples stored as single-use aqueous solutions at +4 degrees C and lyophilized samples stored at +4 and -20 degrees C were the most stable. Samples stored as frozen aqueous solutions at -20 degrees C were the least stable. PMID:19549501

  7. Fabrication of SO/sub 2/preparation system and calibration of PINSTECH sulfur standard for /sup 34/S/sup 32/S mass spectrometric analysis

    International Nuclear Information System (INIS)

    This report describes the fabrication and standardization of operation procedures of a SO/sub 2/ preparation system used for the extraction of sulfur dioxide gas from sulfur minerals (aqueous sulfate, elemental sulfur, and sulfides) for sulfur isotope ratio measurements on a gas source mass spectrometer for hydrological, geological and environmental applications. SO/sub 2/ preparation procedure as described by Fumitaka Yanagisawa and Hitoshi Sakai (1983) is adopted with some modifications. A chemically pure BaSO/sub 4/ powder is chosen as PINSTECH Sulfur Standard PSS-I for routine laboratory /sup 34/S analysis. PSS-1 is calibrated against the International Atomic Energy Agency (IAEA) standard Cannon Diablo Troilite (CDT) using the NBS-127 sulfur standard Sigma /sup 34/S values of PSS-1 as analyzed at PINSTECH and Institute fur Hydorlogie, Munich are found to be 14.58 +-0.07 % CDT (n=6) and 14.59+-0.15% CDT (n=2) respectively. NBS-127 is BaSO/sub 4/ powder from the National Bureau of Standards, USA and has been calibrated against CDT. Interlaboratory comparison of various standards is also documented. Using this system, the reproducibility of sulfur isotope ratio measurements is better than +-0.2 % (n=10). (author)

  8. Air kerma standardization for diagnostic radiology, and requirements proposal for calibration laboratories; Padronizacao da grandeza Kerma no ar para radiodiagnostico e proposta de requisitos para laboratorios de calibracao

    Energy Technology Data Exchange (ETDEWEB)

    Ramos, Manoel Mattos Oliveira

    2009-07-01

    The demand for calibration services and quality control in diagnostic radiology has grown in the country since the publication of the governmental regulation 453, issued by the Ministry of Health in 1998. At that time, to produce results facing the new legislation, many laboratories used different standards and radiation qualities, some of which could be inadequate. The international standards neither supplied consistent radiation qualities and standardization for the different types of equipment available. This situation changed with the publication of the new edition of the IEC 61267 standard, published in 2005. A metrology network was created, but it is not yet accredited by the accreditation organism of the country, INMETRO. The objective of this work was to implement the standardization of the air kerma for the un attenuated qualities (RQR) of IEC 61267, and to develop a requirement proposal for instruments calibration laboratories. Results of interlaboratory comparisons demonstrate that the quantity is standardized and internationally traceable. A laboratory requirement proposal was finalized and it shall be submitted to INMETRO to be used as auxiliary normative document in laboratory accreditation. (author)

  9. Standardization of 64Cu and 68Ga by the 4π(PC)β-γ coincidence method and calibration of the ionization chamber.

    Science.gov (United States)

    Sahagia, M; Luca, A; Antohe, A; Ivan, C

    2012-09-01

    The paper treats the application of the 4π(PC)β-γ coincidence method for the standardization of the radionuclides (64)Cu and (68)Ga. The general coincidence equations are written. Two types of extrapolation were described and used in measurement: the positron-annihilation coincidence, and the counting of all emitted radiations; both methods are compared with respect to results, advantages and drawbacks. The impurities' content correction was applied. The standardized solutions were used to calibrate the ionization chamber CENTRONIC IG12/20A and to determine the gamma-rays emission intensities.

  10. ORNL calibrations facility

    International Nuclear Information System (INIS)

    The ORNL Calibrations Facility is operated by the Instrumentation Group of the Industrial Safety and Applied Health Physics Division. Its primary purpose is to maintain radiation calibration standards for calibration of ORNL health physics instruments and personnel dosimeters. This report includes a discussion of the radioactive sources and ancillary equipment in use and a step-by-step procedure for calibration of those survey instruments and personnel dosimeters in routine use at ORNL

  11. Collaborative Calibration of Control Standards of Hemocoagulase Bothrops Atrox%矛头蝮蛇血凝酶工作标准品的协作标定

    Institute of Scientific and Technical Information of China (English)

    李秀琳; 孙东; 丁忠福; 宁静; 于翀; 崔亮亮; 薛雁

    2016-01-01

    目的:标定我公司生产矛头蝮蛇血凝酶工作标准品。方法:以蛇毒血凝酶质量标准WS-349( X-317)-2001中效价测定为基础,改进测定方法,组织多名实验人员协作标定,并对数据进行统计分析。结果:本批矛头蝮蛇血凝酶工作标准品酶活性单位为103 U/支,加权均值标准差为0.8068,95%加权均值可信限的范围为101.0~105.1 U/支。结论:我公司生产的矛头蝮蛇血凝酶工作标准品标化科学、赋值准确,可以用于本公司产品的效价检测。%Objective:To calibrate the control standards of Hemocoagulase Bothrops atrox of our company.Methods:The enzyme ac-tivity was assayed based on the titration of Hemocoagulase Atrox Quality Standard WS-349(X-317)-2001, the method was modified, more than a number of experimenters were organized to collaboratively calibrate and the data was statistically analyzed.Results: The activity of the control standards of Hemocoagulase Bothrops atrox was 103 units per bottle, the weighted average standard deviation was 0.806 8, and the 95%weighted mean confidence range was 101.0-105.1 units per bottle.Conclusion:The calibration of the control standards of Hemocoagulase Bothrops atrox is scientific and accurate, and can be used to detect the titers of our company′s products .

  12. Calibration of the modulation transfer function of surface profilometers with binary pseudo-random test standards: expanding the application range

    International Nuclear Information System (INIS)

    A modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) gratings and arrays (Proc. SPIE 7077-7 (2007), Opt. Eng. 47, 073602 (2008)) has been proven to be an effective MTF calibration method for a number of interferometric microscopes and a scatterometer (Nucl. Instr. and Meth. A616, 172 (2010)). Here we report on a further expansion of the application range of the method. We describe the MTF calibration of a 6 inch phase shifting Fizeau interferometer. Beyond providing a direct measurement of the interferometer's MTF, tests with a BPR array surface have revealed an asymmetry in the instrument's data processing algorithm that fundamentally limits its bandwidth. Moreover, the tests have illustrated the effects of the instrument's detrending and filtering procedures on power spectral density measurements. The details of the development of a BPR test sample suitable for calibration of scanning and transmission electron microscopes are also presented. Such a test sample is realized as a multilayer structure with the layer thicknesses of two materials corresponding to BPR sequence. The investigations confirm the universal character of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  13. Calibration of the modulation transfer function of surface profilometers with binary pseudo-random test standards: expanding the application range

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy V.; Anderson, Erik H.; Barber, Samuel K.; Bouet, Nathalie; Cambie, Rossana; Conley, Raymond; McKinney, Wayne R.; Takacs, Peter Z.; Voronov, Dmitriy L.

    2011-03-14

    A modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) gratings and arrays [Proc. SPIE 7077-7 (2007), Opt. Eng. 47, 073602 (2008)] has been proven to be an effective MTF calibration method for a number of interferometric microscopes and a scatterometer [Nucl. Instr. and Meth. A616, 172 (2010)]. Here we report on a further expansion of the application range of the method. We describe the MTF calibration of a 6 inch phase shifting Fizeau interferometer. Beyond providing a direct measurement of the interferometer's MTF, tests with a BPR array surface have revealed an asymmetry in the instrument's data processing algorithm that fundamentally limits its bandwidth. Moreover, the tests have illustrated the effects of the instrument's detrending and filtering procedures on power spectral density measurements. The details of the development of a BPR test sample suitable for calibration of scanning and transmission electron microscopes are also presented. Such a test sample is realized as a multilayer structure with the layer thicknesses of two materials corresponding to BPR sequence. The investigations confirm the universal character of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  14. Calibration of the modulation transfer function of surface profilometers with binary pseudo-random test standards: Expanding the application range

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy V; Anderson, Erik H.; Barber, Samuel K.; Bouet, Nathalie; Cambie, Rossana; Conley, Raymond; McKinney, Wayne R.; Takacs, Peter Z.; Voronov, Dmitriy L.

    2010-07-26

    A modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) gratings and arrays [Proc. SPIE 7077-7 (2007), Opt. Eng. 47(7), 073602-1-5 (2008)] has been proven to be an effective MTF calibration method for a number of interferometric microscopes and a scatterometer [Nucl. Instr. and Meth. A 616, 172-82 (2010]. Here we report on a significant expansion of the application range of the method. We describe the MTF calibration of a 6 inch phase shifting Fizeau interferometer. Beyond providing a direct measurement of the interferometer's MTF, tests with a BPR array surface have revealed an asymmetry in the instrument's data processing algorithm that fundamentally limits its bandwidth. Moreover, the tests have illustrated the effects of the instrument's detrending and filtering procedures on power spectral density measurements. The details of the development of a BPR test sample suitable for calibration of scanning and transmission electron microscopes are also presented. Such a test sample is realized as a multilayer structure with the layer thicknesses of two materials corresponding to BPR sequence. The investigations confirm the universal character of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  15. SEALDH-II: An airborne, autonomous, calibration-free TDLAS Hygrometer. First in-flight results with metrological links to the German Primary Humidity Standards

    Science.gov (United States)

    Buchholz, B.; Ebert, V.

    2014-12-01

    The accuracy and comparability of airborne hygrometers remains under debate often only reaching ±10% under static lab conditions or even >±30% in flight. This often limits atmospheric models as well as other species sensors which need to be corrected for water vapor dilution or cross sensitivity. Further challenges come from the huge dynamic range (2-40000 ppmv), the strong spatial gradients (up to several 1000 ppmv/s). and the difficult and error-prone calibration processes, very often without any direct links to the accurate global metrological water vapor scales, which are defined via national primary humidity generators. To fill this gap and provide "traceable", i.e. metrologically validated, airborne hygrometers we developed the Selective Extractive Airborne Laser Diode Hygrometer, SEALDH, which uses dTDLAS with a special, calibration-free data evaluation to circumvent the need for frequent field calibrations and to ensure high accuracy and comparability by a holistic data quality assurance concept in combination with extensive metrological validations at national primary standards. Thus SEALDH-II can be used in a calibration-free field sensor mode (with an absolute, metrologically defined uncertainty of 4.3% +- 3ppmv). The response time is mainly limited by the gas flow and significantly below 1 sec, yielding precision down to 0.08 ppmv (1σ, 1sec) measured at 600 ppmv and 1000 hPa. Its excellent long-term stability, field transfer standard. SEALDH-II has been recently operated without any failures for over 50 hours on several airborne science missions (DENCHAR, AIRTOSS-I, and AIRTOSS-II). In addition SEALDH participated in the representative, international comparison AquaVIT-II. The performance during these missions and the value of a traceability will be demonstrated and discussed.

  16. In-situ calibration of clinical built-in KAP meters with traceability to a primary standard using a reference KAP meter

    Science.gov (United States)

    Malusek, A.; Helmrot, E.; Sandborg, M.; Grindborg, J.-E.; Alm Carlsson, G.

    2014-12-01

    The air kerma-area product (KAP) is used for settings of diagnostic reference levels. The International Atomic Energy Agency (IAEA) recommends that doses in diagnostic radiology (including the KAP values) be estimated with an accuracy of at least ±7% (k = 2). Industry standards defined by the International Electrotechnical Commission (IEC) specify that the uncertainty of KAP meter measurements should be less than +/- 25% (k = 2). Medical physicists willing to comply with the IAEA’s recommendation need to apply correction factors to KAP values reported by x-ray units. The aim of this work is to present and evaluate a calibration method for built-in KAP meters on clinical x-ray units. The method is based on (i) a tandem calibration method, which uses a reference KAP meter calibrated to measure the incident radiation, (ii) measurements using an energy-independent ionization chamber to correct for the energy dependence of the reference KAP meter, and (iii) Monte Carlo simulations of the beam quality correction factors that correct for differences between beam qualities at a standard laboratory and the clinic. The method was applied to the KAP meter in a Siemens Aristos FX plus unit. It was found that values reported by the built-in KAP meter differed from the more accurate values measured by the reference KAP meter by more than 25% for high tube voltages (more than 140 kV) and heavily filtered beams (0.3 mm Cu). Associated uncertainties were too high to claim that the IEC’s limit of 25% was exceeded. Nevertheless the differences were high enough to justify the need for a more accurate calibration of built-in KAP meters.

  17. Measuring laser carrier-envelope phase effects in the noble gases with an atomic hydrogen calibration standard

    CERN Document Server

    Khurmi, Champak; U, Satya Sainadh; Ivanov, I A; Kheifets, A S; Tong, X M; Litvinyuk, I V; Sang, R T; Kielpinski, D

    2016-01-01

    We present accurate measurements of carrier-envelope phase effects on ionisation of the noble gases with few-cycle laser pulses. The experimental apparatus is calibrated by using atomic hydrogen data to remove any systematic offsets and thereby obtain accurate CEP data on other generally used noble gases such as Ar, Kr and Xe. Experimental results for H are well supported by exact TDSE theoretical simulations however significant differences are observed in case of noble gases.

  18. (60)Co in cast steel matrix: A European interlaboratory comparison for the characterisation of new activity standards for calibration of gamma-ray spectrometers in metallurgy.

    Science.gov (United States)

    Tzika, Faidra; Burda, Oleksiy; Hult, Mikael; Arnold, Dirk; Marroyo, Belén Caro; Dryák, Pavel; Fazio, Aldo; Ferreux, Laurent; García-Toraño, Eduardo; Javornik, Andrej; Klemola, Seppo; Luca, Aurelian; Moser, Hannah; Nečemer, Marijan; Peyrés, Virginia; Reis, Mario; Silva, Lidia; Šolc, Jaroslav; Svec, Anton; Tyminski, Zbigniew; Vodenik, Branko; Wätjen, Uwe

    2016-08-01

    Two series of activity standards of (60)Co in cast steel matrix, developed for the calibration of gamma-ray spectrometry systems in the metallurgical sector, were characterised using a European interlaboratory comparison among twelve National Metrology Institutes and one international organisation. The first standard, consisting of 14 disc shaped samples, was cast from steel contaminated during production ("originally"), and the second, consisting of 15 similar discs, from artificially-contaminated ("spiked") steel. The reference activity concentrations of (60)Co in the cast steel standards were (1.077±0.019) Bqg(-1) on 1 January 2013 12h00 UT and (1.483±0.022) Bqg(-1) on 1 June 2013 12h00 UT, respectively.

  19. (60)Co in cast steel matrix: A European interlaboratory comparison for the characterisation of new activity standards for calibration of gamma-ray spectrometers in metallurgy.

    Science.gov (United States)

    Tzika, Faidra; Burda, Oleksiy; Hult, Mikael; Arnold, Dirk; Marroyo, Belén Caro; Dryák, Pavel; Fazio, Aldo; Ferreux, Laurent; García-Toraño, Eduardo; Javornik, Andrej; Klemola, Seppo; Luca, Aurelian; Moser, Hannah; Nečemer, Marijan; Peyrés, Virginia; Reis, Mario; Silva, Lidia; Šolc, Jaroslav; Svec, Anton; Tyminski, Zbigniew; Vodenik, Branko; Wätjen, Uwe

    2016-08-01

    Two series of activity standards of (60)Co in cast steel matrix, developed for the calibration of gamma-ray spectrometry systems in the metallurgical sector, were characterised using a European interlaboratory comparison among twelve National Metrology Institutes and one international organisation. The first standard, consisting of 14 disc shaped samples, was cast from steel contaminated during production ("originally"), and the second, consisting of 15 similar discs, from artificially-contaminated ("spiked") steel. The reference activity concentrations of (60)Co in the cast steel standards were (1.077±0.019) Bqg(-1) on 1 January 2013 12h00 UT and (1.483±0.022) Bqg(-1) on 1 June 2013 12h00 UT, respectively. PMID:27236833

  20. X射线环境水平标准装置的研究与建立%The establishment of x-ray environmental standard calibration quality

    Institute of Scientific and Technical Information of China (English)

    欧向明; 赵士庵; 丁艳秋

    2014-01-01

    Objective:According to the requirement of ISO and National Standard, new X-ray environmental standard for the calibration of environmental dosimeter was established in Beijing SSDL. Methods:Half value layer and air kerma dose-rate was determined in the X-ray environmental qualities by a standard dosimeter. Results: The technical requirement of ISO and National Standard are met within 2% for the X-ray environmental standard qualities in Beijing SSDL. Conclusion:The X-ray standard radiation field can be used for the calibration of environmental dosimeters and experiment of research.%目的:按照国际标准化组织(ISO)技术规范及我国现行国家标准的相关要求,在国家二级标准剂量学实验室建立用于校准X射线环境水平剂量仪的标准装置。方法:通过标准剂量仪测量不同附加过滤后的X射线束的衰减,测定标准辐射场常用的半值层及空气比释动能率,确定环境水平X射线辐射线质的有效能量和平均能量。结果:测定的辐射线质等参数与ISO和国家标准给出的技术指标误差<2%,满足标准装置的要求。结论:建立的X射线环境水平标准装置可以用于校准环境水平辐射剂量仪、并为开展相关的研究工作提供实验条件。

  1. Critical issues for implementation of the standard NBR ISO/IEC 17025:2005 in Testing and Calibration Laboratory: case study at a public institution

    International Nuclear Information System (INIS)

    The public institution aims to promote excellence in public management to contribute to the quality of services provided to its customers and to increase competitiveness in the country, as well as its international projection. A technical barrier to trade that can lead to dissatisfaction and achieve the reputation of the institution is failing the test or calibration results and measurement data, thereby accreditation is regarded as the first essential step to facilitate the mutual acceptance of test results and calibration or measurement data. For recognition, laboratories need to demonstrate full compliance with both the sections of ISO/IEC 17025:2005, i.e. management and technical requirements. This research aims to discuss the critical aspects for implementation of ABNT NBR ISO / IEC 17025:2005 for calibration and testing of a Public Institution seeking accreditation of its laboratories with INMETRO, national accreditation body Laboratories. Besides getting preventive, corrective and improvement actions continues guidelines. Furthermore, the methodology used was to conduct a literature search and apply a questionnaire to identify the degree of agreement / disagreement of the foundations of the standard servers. Analysis of the results showed that the critical issues were: commitment, training, resources (infrastructure, human) and culture. (author)

  2. Fabrication of grating-like polystyrene latex monolayer structure as three-dimensional calibration standards for scanning probe microscope

    Institute of Scientific and Technical Information of China (English)

    Zhu Guo-Dong; Zeng Zhi-Gang; Guo Zhang; Du Qiang-Guo; Yan Xue-Jian

    2009-01-01

    This paper illuminates the preparation of grating-like polystyrene latex monolayer structure, which can minimize the effects of the size deviation of spheres and the defect transfer on the accuracy as calibration samples for micro-scopes. The latex films are grown on freshly cleaved mica substrates by vertical deposition method. The concentration dependence of the structure and the topography of latex films is characterized by optical microscope, ultraviolet-visible transmission spectrum and scanning probe microscope. The origination of such a grating-like structure is also discussed.

  3. Electroluminescent TCC, C3dg and fB/Bb epitope assays for profiling complement cascade activation in vitro using an activated complement serum calibration standard.

    Science.gov (United States)

    van Vuuren, B Jansen; Bergseth, G; Mollnes, T E; Shaw, A M

    2014-01-15

    Electroluminescent assays for epitopes on the complement components C3dg, terminal complement complex (TCC) and factor B/Bb (fB/Bb) have been developed with capture and detection antibodies to produce detection limits C3dg=91±9ng/mL, TCC=3±0.1ng/mL and fB=55.7±0.1ng/mL. The assay performance was assessed against a series of zymosan and heat aggregated IgG (HAIgG) in vitro activations of complement using a calibrated activated complement serum (ACS) as calibration standard. The ACS standard was stable within 20% accuracy over a 6-month period with freeze-thaw cycles as required. Differential activation of the complement cascade was observed for TCC showing a pseudo-first order formation half-life of 3.5h after activation with zymosan. The C3dg activation fragment indicates a 10% total activation for both activation agents. The kinetic-epitope analysis for fB indicates that the capture epitope is on the fB/Bb protein fragment which can then become covered by the formation of C3bBb or C3bBbP complexes during the time course of the cascade.

  4. Report on key comparison COOMET.AUV.A-K5: pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 kHz

    Science.gov (United States)

    Dobrowolska, D.; Kosterov, A.

    2016-01-01

    This is the final report for regional key comparison COOMET.AUV.A-K5 on the pressure calibration of laboratory standard microphones in the frequency range from 2 Hz to 10 kHz. Two laboratories—Central Office of Measures (GUM)—the national metrology institute for Poland and the State Enterprise Scientific-Research Institute for Metrology of Measurement and Control Systems (DP NDI Systema)— the designated institute for acoustics in Ukraine took part in this comparison with the GUM as a pilot. One travelling type LS1P microphone was circulated to the participants and results in the form of regular calibration certificates were collected. The results of the DP NDI Systema obtained in this comparison were linked to the CCAUV.A-K5 key comparison through the joint participation of the GUM. The degrees of equivalence were computed for DP NDI Systema with respect to the CCAUV.A-K5 key comparison reference value. Main text To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  5. Preparation of Calibration Standards of N1-H Paralytic Shellfish Toxin Analogues by Large-Scale Culture of Cyanobacterium Anabaena circinalis (TA04

    Directory of Open Access Journals (Sweden)

    Toshiyuki Suzuki

    2011-03-01

    Full Text Available Mouse bioassay is the official testing method to quantify paralytic shellfish toxins (PSTs in bivalves. A number of alternative analytical methods have been reported. Some methods have been evaluated by a single laboratory validation. Among the different types of methods, chemical analyses are capable of identifying and quantifying the toxins, however a shortage of the necessary calibration standards hampers implementation of the chemical analyses in routine monitoring of PSTs in bivalves. In our present study, we studied preparation of major PST analogues as calibrants by large-scale cultivation of toxic freshwater cyanobacteria Anabaena circinalis TA04. The cells were steadily grown in 10 L bottle for 28 days. The primary N1-H toxins, C1/C2, were produced at a concentration of 1.3 ± 0.1 µmol/L. The intracellular and extracellular toxins occupied 80% and 20%, respectively. Over 220 µmol of the toxins was obtained from approximately 200 L of the culture over six months, demonstrating that it is sufficient to prepare saxitoxin analogues. The toxins were chemically converted to six N1-H analogues. Preparation of the analogues was carried out at relatively high yields (50–90%. The results indicate that our preparation method is useful to produce N1-H toxins. In our present study, detailed conditions for preparation of one of the rare N1-H analogues, gonyautoxin-5, were investigated.

  6. The energy calibration and precision of a gamma spectrometry unit - Method using the electron annihilation energy as the only standard

    International Nuclear Information System (INIS)

    Spectrometry using Ge(Li) detectors is discussed. The excellent resolution of this type of detector, the mathematical analysis of the spectral lines of the pulses, and the reproducibility of the spectrometer enable highly accurate measurements of the abscises (some 10-5) corresponding to the peaks. A method using the annihilation energy of the electron as the only standard was developed. The method is applied to the measurement of the gamma ray energies of the radioelements: 22Na, 24Na, 56Mn, 56Co, 59Fe, 72Ga, 88Y, 122Sb, 124Sb and 137Cs. (author)

  7. The PROMIS Physical Function item bank was calibrated to a standardized metric and shown to improve measurement efficiency

    DEFF Research Database (Denmark)

    Rose, Matthias; Bjørner, Jakob; Gandek, Barbara;

    2014-01-01

    of 16,065 adults answered item subsets (n>2,200/item) on the Internet, with oversampling of the chronically ill. Classical test and item response theory methods were used to evaluate 149 PROMIS PF items plus 10 Short Form-36 and 20 Health Assessment Questionnaire-Disability Index items. A graded....... In simulations, a 10-item computerized adaptive test (CAT) eliminated floor and decreased ceiling effects, achieving higher measurement precision than any comparable length static tool across four SDs of the measurement range. Improved psychometric properties were transferred to the CAT's superior ability...... to identify differences between age and disease groups. CONCLUSION: The item bank provides a common metric and can improve the measurement of PF by facilitating the standardization of patient-reported outcome measures and implementation of CATs for more efficient PF assessments over a larger range....

  8. Sandia WIPP calibration traceability

    International Nuclear Information System (INIS)

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities

  9. Calibration of farmer dosemeters

    International Nuclear Information System (INIS)

    The Farmer Dosemeters of Atomic Energy Medical Centre (AEMC) Jamshoro were calibrated in the Secondary Standard Dosimetry Laboratory (SSDL) at PINSTECH, using the NPL Secondary Standard Therapy level X-ray exposure meter. The results are presented in this report. (authors)

  10. Automatic Titrator Calibration Standard Solution of Silver Nitrate%自动电位滴定仪标定硝酸银标准溶液

    Institute of Scientific and Technical Information of China (English)

    马玉芬

    2012-01-01

      Using the configuration DM141-SC-type silver electrode DL58 automatic titrator calibration standard solution of silver nitrate, accuracy and precision compared with manual titration with a large degree of increase, and simple operation, equipment maintenance, more accurate titration results .Titration process signal acquisition is 0.1mV, minimum feed rate can be titrated to achieve 0.001ml, is the recommended method.%  采用配置DM141-SC型银电极的DL58型自动电位滴定仪标定硝酸银标准溶液,准确度及精密度均较手动滴定有很大程度的提高,且操作简单,仪器维护方便,滴定结果更准确,滴定过程采集信号为0.1mV,滴定最小进给量可达到0.001mL,是目前值得推荐的方法.

  11. Noble metals as permanent chemical modifiers for the determination of mercury in environmental reference materials using solid sampling graphite furnace atomic absorption spectrometry and calibration against aqueous standards

    Science.gov (United States)

    da Silva, Alessandra Furtado; Welz, Bernhard; Curtius, Adilson J.

    2002-12-01

    Iridium, palladium, rhodium and ruthenium, thermally deposited on the platform, were investigated as permanent modifiers for the determination of mercury in ash, sludge, marine and river sediment reference materials, ground to a particle size of 50 μm, using solid sampling graphite furnace atomic absorption spectrometry. A total mass of 250 μg of each modifier was applied using 25 injections of 20 μl of modifier solution (500 mg l -1), and executing a temperature program for modifier conditioning after each injection. The performance of palladium was found to be most consistent, taking the characteristic mass as the major criterion, resulting in an excellent correlation between the measured integrated absorbance values and the certified mercury contents. Mercury was found to be lost in part from aqueous solutions during the drying stage in the presence of all the investigated permanent modifiers, as well as in the presence of the palladium and magnesium nitrates modifier added in solution. A loss-free determination of mercury in aqueous solutions could be reached only after the addition of potassium permanganate, which finally made possible the use of aqueous standards for the direct analysis of solid samples. A characteristic mass of 55-60 pg Hg was obtained for the solid samples, using Pd as a permanent modifier, and also in aqueous solutions after the addition of permanganate. The results obtained for mercury in ash, sludge and sediment reference materials, using direct solid sapling and calibration against aqueous standards, as well as the detection limit of 0.2 mg kg -1 were satisfactory for a routine procedure.

  12. Noble metals as permanent chemical modifiers for the determination of mercury in environmental reference materials using solid sampling graphite furnace atomic absorption spectrometry and calibration against aqueous standards

    Energy Technology Data Exchange (ETDEWEB)

    Silva, Alessandra Furtado da; Welz, Bernhard; Curtius, Adilson J

    2002-12-02

    Iridium, palladium, rhodium and ruthenium, thermally deposited on the platform, were investigated as permanent modifiers for the determination of mercury in ash, sludge, marine and river sediment reference materials, ground to a particle size of 50 {mu}m, using solid sampling graphite furnace atomic absorption spectrometry. A total mass of 250 {mu}g of each modifier was applied using 25 injections of 20 {mu}l of modifier solution (500 mg l{sup -1}), and executing a temperature program for modifier conditioning after each injection. The performance of palladium was found to be most consistent, taking the characteristic mass as the major criterion, resulting in an excellent correlation between the measured integrated absorbance values and the certified mercury contents. Mercury was found to be lost in part from aqueous solutions during the drying stage in the presence of all the investigated permanent modifiers, as well as in the presence of the palladium and magnesium nitrates modifier added in solution. A loss-free determination of mercury in aqueous solutions could be reached only after the addition of potassium permanganate, which finally made possible the use of aqueous standards for the direct analysis of solid samples. A characteristic mass of 55-60 pg Hg was obtained for the solid samples, using Pd as a permanent modifier, and also in aqueous solutions after the addition of permanganate. The results obtained for mercury in ash, sludge and sediment reference materials, using direct solid sapling and calibration against aqueous standards, as well as the detection limit of 0.2 mg kg{sup -1} were satisfactory for a routine procedure.

  13. The influence of the calibration standard and the chemical composition of the water samples residue in the counting efficiency of proportional detectors for gross alpha and beta counting. Application on the radiologic control of the IPEN-CNEN/SP

    International Nuclear Information System (INIS)

    In this work the efficiency calibration curves of thin-window and low background gas-flow proportional counters were determined for calibration standards with different energies and different absorber thicknesses. For the gross alpha counting we have used 241Am and natural uranium standards and for the gross beta counting we have used 90Sr/90Y and 137Cs standards in residue thicknesses ranging from 0 to approximately 18 mg/cm2. These sample thicknesses were increased with a previously determined salted solution prepared simulating the chemical composition of the underground water of IPEN The counting efficiency for alpha emitters ranged from 0,273 +- 0,038 for a weightless residue to only 0,015 +- 0,002 in a planchet containing 15 mg/cm2 of residue for 241Am standard. For natural uranium standard the efficiency ranged from 0,322 +- 0,030 for a weightless residue to 0,023 +- 0,003 in a planchet containing 14,5 mg/cm2 of residue. The counting efficiency for beta emitters ranged from 0,430 +- 0,036 for a weightless residue to 0,247 +- 0,020 in a planchet containing 17 mg/cm2 of residue for 137Cs standard. For 90Sr/90Y standard the efficiency ranged from 0,489 +- 0,041 for a weightless residue to 0,323 +- 0,026 in a planchet containing 18 mg/cm2 of residue. Results make evident the counting efficiency variation with the alpha or beta emitters energies and the thickness of the water samples residue. So, the calibration standard, the thickness and the chemical composition of the residue must always be considered in the gross alpha and beta radioactivity determination in water samples. (author)

  14. Primary Research on the Standard System of Air Vehicle Calibration Models Used in Wind Tunnel Test%飞行器风洞试验标模体系研究初探

    Institute of Scientific and Technical Information of China (English)

    战培国; 罗月培

    2011-01-01

    风洞试验标模是一种评估风洞试验准度和验证CFD算法的通用校准模型.本文归纳分析了北大西洋公约组织AGARD系列、法国ONERAM系列为代表的国外风洞试验标模发展情况,阐述了我国DBM、GBM、HSCM风洞标模系列的发展现状,探讨了建立和完善风洞试验标模体系的一些问题,意在为国内风洞和试验技术发展提供参考.%The standard air vehicle calibration model is a common calibration model used for evaluating wind tunnel test results and validating CFD methods.This paper analyses the development of overseas wind tunnel calibration models,typically,NATO AGARD series and ONERA M series in France,introduces state of the art domestic calibration models,such as DBM,CBM and HSCM series,discusses some problems about setting up and consummating the system of calibration models.The purpose is to provide thoughts and references for the development of domestic wind tunnel techniques.

  15. 基于ASAM标准的汽车电控系统匹配标定系统设计%Design of Automotive Electronic Control System Calibration System based on ASAM Standards

    Institute of Scientific and Technical Information of China (English)

    温泉; 张广秀; 张建

    2012-01-01

    ASAM-MCD standards is a set of widely-agreed calibration criterion in industry. Application of the criterion can standardize data exchange methods, software and hardware tools development, and thereby decrease development cost and periods. This paper discusses the main features of CCP and ASAP2 standardized interface for the description data. We have developed the electric vehicle control unit calibration system according to the ASAM-MCD standards. The system can implement file-based parameters monitoring, measurement and calibration.%ASAM—MCD标准是一套得到工业界广泛认可的匹配标定规范。通过该规范的实施,可以对ECU开发过程中用到的数据交换方法和软硬件工具进行标准化,从而减少汽车电子领域的开发成本、缩短开发周期。本文论述了ASAM—MCD标准中的CAN标定协议和ASAP2数据接口规范的主要特点,并根据该标准开发了电动汽车用整车控制器匹配标定系统。通过该系统可以实现基于文件的参数监控、测量和标定。

  16. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  17. Calibration Database Logic Analysis and Implementation based on ASAM Standard%基于ASAM标准的标定数据库逻辑分析与编程实现

    Institute of Scientific and Technical Information of China (English)

    宋维群; 杨世春; 李明; 崔海港; 曹耀光

    2012-01-01

    In this article, ASAM -2MC standard proposed by ASAM organization standardizes data needed in calibration of automotive electronic control system. In-depth analysis is made to ASAM-2MC standard and logic relation and database algorithm between calibrated information items are introduced, and calibration database management software in compliance with ASAM-2MC standard is developed. The interpretation of a21 description document used in calibration of engine electronic control system shows that this software is able to classify control parameters in vehicle development, display and provide visual interface to analyze and correct the items in the interface.%ASAM组织提出的ASAM-2MC标准对汽车电子控制系统标定过程中所需要的数据信息进行了规范.深入分析了ASAM-2MC标准并提出了相应的标定信息项目实体之间的逻辑关系及数据库解析算法,开发了符合ASAM-2MC标准的标定数据库管理软件.通过对发动机电控系统标定过程中使用的a21描述文件的解析表明,该软件能够对整车开发过程中与控制相关的参数信息进行分类,显示并提供可视化界面对其中的项目进行分析与修改.

  18. NVLAP calibration laboratory program

    Energy Technology Data Exchange (ETDEWEB)

    Cigler, J.L.

    1993-12-31

    This paper presents an overview of the progress up to April 1993 in the development of the Calibration Laboratories Accreditation Program within the framework of the National Voluntary Laboratory Accreditation Program (NVLAP) at the National Institute of Standards and Technology (NIST).

  19. NHEXAS PHASE I ARIZONA STUDY--STANDARD OPERATING PROCEDURE FOR OPERATION, CALIBRATION AND MAINTENANCE OF THE SENTEX SCENTOGUN PORTABLE PHOTOIONIZATION DETECTOR (UA-F-4.1)

    Science.gov (United States)

    The purpose of this SOP is to describe the procedures for the operation, calibration, and maintenance of the Sentex Scentogun portable photoionization detector (PID). The sampling Scentogun was used to collect volatile organic compound (VOC) levels during the Arizona NHEXAS proj...

  20. Strain Gauge Balance Calibration and Data Reduction at NASA Langley Research Center

    Science.gov (United States)

    Ferris, A. T. Judy

    1999-01-01

    This paper will cover the standard force balance calibration and data reduction techniques used at Langley Research Center. It will cover balance axes definition, balance type, calibration instrumentation, traceability of standards to NIST, calibration loading procedures, balance calibration mathematical model, calibration data reduction techniques, balance accuracy reporting, and calibration frequency.

  1. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John F. Schabron; Joseph F. Rovani; Susan S. Sorini

    2007-03-31

    The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005, requires that calibration of mercury continuous emissions monitors (CEMs) be performed with NIST-traceable standards. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The traceability protocol will be written by EPA. Traceability will be based on the actual analysis of the output of each calibration unit at several concentration levels ranging from about 2-40 ug/m{sup 3}, and this analysis will be directly traceable to analyses by NIST using isotope dilution inductively coupled plasma/mass spectrometry (ID ICP/MS) through a chain of analyses linking the calibration unit in the power plant to the NIST ID ICP/MS. Prior to this project, NIST did not provide a recommended mercury vapor pressure equation or list mercury vapor pressure in its vapor pressure database. The NIST Physical and Chemical Properties Division in Boulder, Colorado was subcontracted under this project to study the issue in detail and to recommend a mercury vapor pressure equation that the vendors of mercury vapor pressure calibration units can use to calculate the elemental mercury vapor concentration in an equilibrium chamber at a particular temperature. As part of this study, a preliminary evaluation of calibration units from five vendors was made. The work was performed by NIST in Gaithersburg, MD and Joe Rovani from WRI who traveled to NIST as a Visiting Scientist.

  2. Direct chlorine determination in crude oils by energy dispersive X-ray fluorescence spectrometry: An improved method based on a proper strategy for sample homogenization and calibration with inorganic standards

    International Nuclear Information System (INIS)

    Official guidelines to perform chlorine determination in crude oil are (i) American Society for Testing and Materials (ASTM) D6470, which is based on the extraction of water from the oil and subsequent determination of the chloride by potentiometry, (ii) ASTM D3230, that measures the conductivity of a solution of crude oil in a mixture of organic solvents and (iii) US Environmental Protection Agency (EPA) 9075 that uses energy dispersive X-ray fluorescence spectrometry to quantify chlorine and it is applicable for the range from 200 μg g-1 to percent levels of the analyte. The goal of this work is to propose method to quantify lower amounts of chlorine in crude oil using energy dispersive X-ray fluorescence spectrometry using a simple calibration strategy. Sample homogenization procedure was carefully studied in order to enable accurate results. The calibration curve was made with standards prepared by diluting aqueous NaCl standard in glycerin. The method presented a linear response that covers the range from 8 to at least 100 μg g-1 of chlorine. Chlorine in crude oil samples from Campos Basin - Brazil were quantified by the proposed method and by potentiometry after extraction of chlorine from the oil. Results achieved using both methods were statistically the same at 95% confidence level.

  3. Direct chlorine determination in crude oils by energy dispersive X-ray fluorescence spectrometry: An improved method based on a proper strategy for sample homogenization and calibration with inorganic standards

    Science.gov (United States)

    Doyle, Adriana; Saavedra, Alvaro; Tristão, Maria Luiza B.; Nele, Márcio; Aucélio, Ricardo Q.

    2011-05-01

    Official guidelines to perform chlorine determination in crude oil are (i) American Society for Testing and Materials (ASTM) D6470, which is based on the extraction of water from the oil and subsequent determination of the chloride by potentiometry, (ii) ASTM D3230, that measures the conductivity of a solution of crude oil in a mixture of organic solvents and (iii) US Environmental Protection Agency (EPA) 9075 that uses energy dispersive X-ray fluorescence spectrometry to quantify chlorine and it is applicable for the range from 200 μg g - 1 to percent levels of the analyte. The goal of this work is to propose method to quantify lower amounts of chlorine in crude oil using energy dispersive X-ray fluorescence spectrometry using a simple calibration strategy. Sample homogenization procedure was carefully studied in order to enable accurate results. The calibration curve was made with standards prepared by diluting aqueous NaCl standard in glycerin. The method presented a linear response that covers the range from 8 to at least 100 μg g - 1 of chlorine. Chlorine in crude oil samples from Campos Basin - Brazil were quantified by the proposed method and by potentiometry after extraction of chlorine from the oil. Results achieved using both methods were statistically the same at 95% confidence level.

  4. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  5. Calibration of the solar radio spectrometer

    Institute of Scientific and Technical Information of China (English)

    2009-01-01

    This paper shows some improvements and new results of calibration of Chinese solar radio spectrometer by analyzing the daily calibration data recorded in the period of 1997-2007. First, the calibration coefficient is fitted for three bands (1.0-2.0 GHz, 2.6-3.8 GHz, 5.2-7.6 GHz) of the spectrometer by using the moving-average method confined by the property of the daily calibration data. By this calibration coefficient, the standard deviation of the calibration result was less than 10 sfu for 95% frequencies of 2.6-3.8 GHz band in 2003. This result is better than that calibrated with the constant coefficient. Second, the calibration coefficient is found in good correlation with local air temperature for most frequencies of 2.6-3.8 GHz band. Moreover, these results are helpful in the research of the quiet solar radio emission.

  6. Calibration of the solar radio spectrometer

    Institute of Scientific and Technical Information of China (English)

    TAN ChengMing; YAN YiHua; TAN BaoLin; XU GuiRong

    2009-01-01

    This paper shows some improvements and new results of calibration of Chinese solar radio spectrom-eter by analyzing the daily calibration data recorded in the period of 1997-2007. First, the calibration coefficient is fitted for three bands (1.0-2.0 GHz, 2.6-3.8 GHz, 5.2-7.6 GHz) of the spectrometer by using the moving-average method confined by the property of the daily calibration data. By this calibration coefficient, the standard deviation of the calibration result was less than 10 sfu for 95% frequencies of 2.6-3.8 GHz band in 2003. This result is better than that calibrated with the constant coefficient. Second, the calibration coefficient is found in good correlation with local air temperature for most frequencies of 2.6-3.8 GHz band. Moreover, these results are helpful in the research of the quiet solar radio emission.

  7. Quantification of technetium-99m hexamethylpropylene amine oxime brain uptake in routine clinical practice using calibrated point sources as an external standard: phantom and human studies

    International Nuclear Information System (INIS)

    Quantitative methods for calculation of regional cerebral blood flow with technetium-99m hexamethyl-propylene amine oxime (99mTc-HMPAO) have been proposed. These methods are very labour intensive and therefore are not useful in routine clinical practice. We describe a simple alternative method, using calibrated point sources as a scaling factor, whereby the tomographic slices are displayed as regional 99mTc-HMPAO brain uptake per cm3 brain tissue in percent of the injected lipophilic dose. The method was validated on Jaszczak and Hoffman phantoms using a three-detector system with HR parallel and HR fan-beam collimators. Under the optimal conditions described in this paper, the measured to real activity ratio was 1.00 (SD=0.06). The reproducibility of the cerebellar uptake in a group of ten normal volunteers and five patients was studied. Intra-individually a mean deviation of 12.6% was observed for the total group. For those patients with a heart rate difference of less than 5 units between the two studies, a mean deviation of 7.2% was obtained. Quantitative 99mTc-HMPAO brain uptake images can be useful for longitudinal studies, especially for follow-up, activation and pharmacological studies. (orig.)

  8. Calibrating Fundamental British Values: How Head Teachers Are Approaching Appraisal in the Light of the Teachers' Standards 2012, Prevent and the Counter-Terrorism and Security Act, 2015

    Science.gov (United States)

    Revell, Lynn; Bryan, Hazel

    2016-01-01

    In requiring that teachers should "not undermine fundamental British values (FBV)," a phrase originally articulated in the Home Office counter-terrorism document, Prevent, the Teachers' Standards has brought into focus the nature of teacher professionalism. Teachers in England are now required to promote FBV within and outside school,…

  9. Absolute sensitivity calibration of extreme ultraviolet photoresists

    Energy Technology Data Exchange (ETDEWEB)

    Jones, Juanita; Naulleau, Patrick P.; Gullikson, Eric M.; Aquila, Andrew; George, Simi; Niakoula, Dimitra

    2008-05-16

    One of the major challenges facing the commercialization of extreme ultraviolet (EUV) lithography remains simultaneously achieving resist sensitivity, line-edge roughness, and resolution requirement. Sensitivity is of particular concern owing to its direct impact on source power requirements. Most current EUV exposure tools have been calibrated against a resist standard with the actual calibration of the standard resist dating back to EUV exposures at Sandia National Laboratories in the mid 1990s. Here they report on an independent sensitivity calibration of two baseline resists from the SEMATECH Berkeley MET tool performed at the Advanced Light Source Calibrations and Standards beamline. The results show the baseline resists to be approximately 1.9 times faster than previously thought based on calibration against the long standing resist standard.

  10. Final report on supplementary comparison of the calibration of a 5 kg stainless steel standard weight between INDECOPI-PERÚ and CEM-ESPAÑA (SIM.M.M-S10)

    Science.gov (United States)

    Quiroga Rojas, Aldo; Nieves Medina, María

    2012-01-01

    This report summarizes the results of the supplementary comparison SIM.M.M-S10 of a 5 kg stainless steel mass standard between INDECOPI (Peru) and CEM (Spain).The objective of the comparison was to demonstrate the metrological equivalence between the two laboratories. The results of the comparison will be used to support the Calibration and Measurement Capabilities (CMCs) declared by the institutes at 5 kg. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the SIM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  11. Dosimetry standards

    International Nuclear Information System (INIS)

    The following leaflets are contained in this folder concerning the National Physical Laboratory's measurement services available in relation to dosimetry standards: Primary standards of X-ray exposure and X-ray irradiation facilities, X-ray dosimetry at therapy levels, Protection-level X-ray calibrations, Therapy-level gamma-ray facility, Fricke dosemeter reference service, Low-dose-rate gamma-ray facility, Penetrameter and kV meter calibration, Measurement services for radiation processing, Dichromate dosemeter reference service, Electron linear accelerator. (U.K.)

  12. Calibration of Nacelle-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  13. Synthetic multi-element standards: a good tool for calibration and quality control of irradiation facilities used for neutron activation analysis

    International Nuclear Information System (INIS)

    Neutron activation analysis (NAA) is a physical technique used for the absolute measurement of the concentration of substances in solids and liquids. The method uses neutron irradiation which is commonly realised using a nuclear reactor in order to activate (make radioactive) different isotopes of the elements present in the sample. The radionuclides produced in this way emit gamma-rays that are characteristic of the elements present in the sample. Using gamma-ray spectrometry these radionuclides can then be identified and quantified, and hence their concentration in the sample can be determined. Although NAA is a straightforward method it requires a sound control of the many physical parameters involved to obtain accurate results and to guarantee a set accuracy in routine analysis. The accuracy of NAA depends on the specific measurement method used. One can perform NAA in a relative way by co-irradiating a known standard and the unknown sample in the same conditions and by comparing the ratio of gamma-rays they emit. Relative NAA has limited applicability since it requires reference standards with a comparable composition as the unknown. A more generally applicable method is the k0-NAA method. In the k0-NAA method all measurements are relative to the element Au resulting in 198Au when irradiated. The k0-NAA method further relies on the fact that the neutron energy spectrum produced in a given position in the reactor can be parameterised with two parameters: the shape factor of the epithermal neutron flux, indicating the deviation of the epithermal neutron spectrum from the ideal 1/E shape approximated by a 1/E1+a distribution, with E the neutron energy; f: the thermal-to-epithermal neutron flux ratio. The parameters f and a are characteristic for the irradiation facility (reactor and irradiation channels) and may change or fluctuate in time according to the irradiation conditions. The way elements activate (become radioactive) when interacting with neutrons is

  14. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  15. Recommended Inorganic Chemicals for Calibration.

    Science.gov (United States)

    Moody, John R.; And Others

    1988-01-01

    All analytical techniques depend on the use of calibration chemicals to relate analyte concentration to instrumental parameters. Discusses the preparation of standard solutions and provides a critical evaluation of available materials. Lists elements by group and discusses the purity and uses of each. (MVL)

  16. Evaluation of the energy dependence of ionization chambers pencil type calibrated beam tomography standards; Avaliacao da dependencia energetica de camaras de ionizacao do tipo lapis calibradas em feixes padroes de tomografia

    Energy Technology Data Exchange (ETDEWEB)

    Fontes, Ladyjane Pereira; Potiens, Maria da Penha A., E-mail: lpfontes@ipen.br [Instituto de Pesquisas Energeticas e Nucleres (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2015-07-01

    The Instrument Calibration Laboratory of IPEN (LCI - IPEN) performs calibrations of pencil-type ionization chambers (IC) used in measures of dosimetric survey on clinical systems of Computed Tomography (CT). Many users make mistakes when using a calibrated ionization chamber in their CT dosimetry systems. In this work a methodology for determination of factors of correction for quality (Kq) through the calibration curve that is specific for each ionization chamber was established. Furthermore, it was possible to demonstrate the energy dependence on an pencil-type Ionization Chamber(IC) calibrated at the LCI - IPEN. (author)

  17. 14 CFR 33.45 - Calibration tests.

    Science.gov (United States)

    2010-01-01

    ... Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF TRANSPORTATION AIRCRAFT AIRWORTHINESS STANDARDS: AIRCRAFT ENGINES Block Tests; Reciprocating Aircraft Engines § 33.45 Calibration tests. (a) Each engine must be subjected to the calibration tests necessary to establish its power characteristics...

  18. Efficiency calibration of low background gamma spectrometer

    International Nuclear Information System (INIS)

    A method of efficiency calibration is described. The authors used standard ores of U, Ra and Th (power form), KCl and Cs-137 sources to do calibration volume-sources which were directly placed on the detector end cap. In such a measuring geometry, it is not necessary to make coincidence-summing correction. The efficiency calibration curve obtained by the method were compared with results measured by Am-241, Cd-109 and Eu-152 calibration sources. The agree in the error of about 5%

  19. Calibration metrology for fixed irradiation sensors

    International Nuclear Information System (INIS)

    After having recalled the regulatory and technical framework of the calibration of radioprotection measurement instruments, and outlined some technical and operational constraints, the authors report the development of an in situ calibration methodology, i.e. without displacement of the sensor. After the presentation of the calibration chain (from the measurement given by a fixed sensor to the reference value given by a primary standard), they indicate the definition and calculation of the different calibration coefficients allowing the linking up of the different levels, and also the taking of uncertainties into account. They finally report the validation of results

  20. Calibration services for medical applications of radiation

    Energy Technology Data Exchange (ETDEWEB)

    DeWerd, L.A.

    1993-12-31

    Calibration services for the medical community applications of radiation involve measuring radiation precisely and having traceability to the National Institute of Standards and Technology (NIST). Radiation therapy applications involve the use of ionization chambers and electrometers for external beams and well-type ionization chamber systems as well as radioactive sources for brachytherapy. Diagnostic x-ray applications involve ionization chamber systems and devices to measure other parameters of the x-ray machine, such as non-invasive kVp meters. Calibration laboratories have been established to provide radiation calibration services while maintaining traceability to NIST. New radiation applications of the medical community spur investigation to provide the future calibration needs.

  1. Tank calibration; Arqueacao de tanques

    Energy Technology Data Exchange (ETDEWEB)

    Chan, Ana [PETROBRAS S.A., Rio de Janeiro, RJ (Brazil)

    2003-07-01

    This work relates the analysis of the norms ISO (International Organization for Standardization) for calibration of vertical cylindrical tanks used in fiscal measurement, established on Joint Regulation no 1 of June 19, 2000 between the ANP (National Agency of Petroleum) and the INMETRO (National Institute of Metrology, Normalization and Industrial Quality). In this work a comparison between norms ISO and norms published by the API (American Petroleum Institute) and the IP (Institute of Petroleum) up to 2001 was made. It was concluded that norms ISO are wider than norms API, IP, and INMETRO methods in the calibration of vertical cylindrical tanks. (author)

  2. SNR 1E 0102.2-7219 as an X-ray Calibration Standard in the 0.5-1.0 keV Bandpass and Its Application to the CCD Instruments aboard Chandra, Suzaku, Swift and XMM-Newton

    CERN Document Server

    Plucinsky, Paul P; Foster, Adam; Haberl, Frank; Miller, Eric D; Pollock, A M T; Sembay, Steve

    2016-01-01

    We desire a simple comparison of the absolute effective areas of the current generation of CCD instruments onboard the following observatories: Chandra ACIS-S3, XMM-Newton (EPIC-MOS and EPIC-pn), Suzaku XIS, and Swift XRT and a straightforward comparison of the time-dependent response of these instruments across their respective mission lifetimes. We have been using 1E 0102.2-7219, the brightest supernova remnant in the Small Magellanic Cloud, to evaluate and modify the response models of these instruments. 1E 0102.2-7219 has strong lines of O, Ne, and Mg below 1.5 keV and little or no Fe emission to complicate the spectrum. As part of the activities of the International Astronomical Consortium for High Energy Calibration (IACHEC), we have developed a standard spectral model for 1E 0102.2-7219. The model is empirical in that it includes Gaussians for the identified lines, an absorption component in the Galaxy, another absorption component in the SMC, and two thermal continuum components. In our fits, the mode...

  3. The CIEMAT Neutron Standards Laboratory (LPN): National Reference Facility for the Calibration of Neutron Measurement Equipment; El Laboratorio de Patrones Neutronicos del CIEMAT: instalacion de referencia nacional para la calibracion de equipos de medida neutronica

    Energy Technology Data Exchange (ETDEWEB)

    Mendez, R.

    2014-04-01

    This is a new facility that, in keeping with the recommendations of standard ISO 8529:1, has 20m5-Be and 252Cf neutron sources stored in water inside a 9 m x 7.5 m x 8 m Irradiation Room with 1.25 m concrete walls that is accessed through a 6.5 ton sliding shielded door. These sources are remotely handled from the Control Room by means of a system that places them in the geometric center of the room and aligns them with the equipment or materials to be irradiated arranged on a base plate. The facility has a series of interlocks that guarantee safety during operations with sources, and the verification measurements have demonstrated the suitability of the shielding used. The LPN will be the answer to the demand for a national laboratory for neutron measurement equipment calibration and equipment and material irradiation under controlled conditions, and it will position our country at the same level as our European partners. (Author)

  4. TOD to TTP calibration

    Science.gov (United States)

    Bijl, Piet; Reynolds, Joseph P.; Vos, Wouter K.; Hogervorst, Maarten A.; Fanning, Jonathan D.

    2011-05-01

    The TTP (Targeting Task Performance) metric, developed at NVESD, is the current standard US Army model to predict EO/IR Target Acquisition performance. This model however does not have a corresponding lab or field test to empirically assess the performance of a camera system. The TOD (Triangle Orientation Discrimination) method, developed at TNO in The Netherlands, provides such a measurement. In this study, we make a direct comparison between TOD performance for a range of sensors and the extensive historical US observer performance database built to develop and calibrate the TTP metric. The US perception data were collected doing an identification task by military personnel on a standard 12 target, 12 aspect tactical vehicle image set that was processed through simulated sensors for which the most fundamental sensor parameters such as blur, sampling, spatial and temporal noise were varied. In the present study, we measured TOD sensor performance using exactly the same sensors processing a set of TOD triangle test patterns. The study shows that good overall agreement is obtained when the ratio between target characteristic size and TOD test pattern size at threshold equals 6.3. Note that this number is purely based on empirical data without any intermediate modeling. The calibration of the TOD to the TTP is highly beneficial to the sensor modeling and testing community for a variety of reasons. These include: i) a connection between requirement specification and acceptance testing, and ii) a very efficient method to quickly validate or extend the TTP range prediction model to new systems and tasks.

  5. 40 CFR 1065.790 - Mass standards.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 32 2010-07-01 2010-07-01 false Mass standards. 1065.790 Section 1065... ENGINE-TESTING PROCEDURES Engine Fluids, Test Fuels, Analytical Gases and Other Calibration Standards § 1065.790 Mass standards. (a) PM balance calibration weights. Use PM balance calibration weights...

  6. e-Calibrations: using the Internet to deliver calibration services in real time at lower cost

    International Nuclear Information System (INIS)

    The National Institute of Standards and Technology (NIST) is expanding into a new frontier in the delivery of measurement services. The Internet will be employed to provide industry with electronic traceability to national standards. This is a radical departure from the traditional modes of traceability and presents many new challenges. The traditional mail-based calibration service relies on sending artifacts to the user, who then mails them back to NIST for evaluation. The new service will deliver calibration results to the industry customer on-demand, in real-time, at a lower cost. The calibration results can be incorporated rapidly into the production process to ensure the highest quality manufacturing. The service would provide the US radiation processing industry with a direct link to the NIST calibration facilities and its expertise, and provide an interactive feedback process between industrial processing and the national measurement standard. Moreover, an Internet calibration system should contribute to the removal of measurement-related trade barriers

  7. Trinocular Calibration Method Based on Binocular Calibration

    Directory of Open Access Journals (Sweden)

    CAO Dan-Dan

    2012-10-01

    Full Text Available In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error of the global calibration of the two camera pairs in the experiments can be as low as 0.277% and 0.328% respectively. Experiment results show that this method is feasible, simple and effective, and has high precision.

  8. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    Science.gov (United States)

    Cornic, Philippe; Illoul, Cédric; Cheminet, Adam; Le Besnerais, Guy; Champagnat, Frédéric; Le Sant, Yves; Leclaire, Benjamin

    2016-09-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data.

  9. LIDAR Velodyne HDL-64E Calibration Using Pattern Planes

    Directory of Open Access Journals (Sweden)

    Gerardo Atanacio-Jiménez

    2011-11-01

    Full Text Available This work describes a method for calibration of the Velodyne HDL‐64E scanning LIDAR system. The principal contribution was expressed by a pattern calibration signature, the mathematical model and the numerical algorithm for computing the calibration parameters of the LIDAR. In this calibration pattern the main objective is to minimize systematic errors due to geometric calibration factor. It describes an algorithm for solution of the intrinsic and extrinsic parameters. Finally, its uncertainty was calculated from the standard deviation of calibration result errors.

  10. Ørsted Pre-Flight Magnetometer Calibration Mission

    DEFF Research Database (Denmark)

    Risbo, T.; Brauer, Peter; Merayo, José M.G.;

    2003-01-01

    The compact spherical coil (CSC) vector-feedback magnetometer on the Danish circle dividersted geomagnetic mapping satellite underwent extensive calibrations and verifications prior to integration and launch. The theory of the 'thin shell' calibration procedure is introduced. Spherical harmonic...... and the overall calibration results are given. The temperature calibrations are explained and reported on. The overall calibration model standard deviation is about 100 pT rms. Comparisons with the later in-flight calibrations show that, except for the unknown satellite offsets, an agreement within 4 n......T was obtained. Finally an rf interference between the CSC and the Overhauser magnetometer is discussed, which may account for some of this discrepancy....

  11. Sensor Calibration Design Based on D-Optimality Criterion

    Directory of Open Access Journals (Sweden)

    Hajiyev Chingiz

    2016-09-01

    Full Text Available In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM. As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.

  12. The physical principles of XRF calibrations

    International Nuclear Information System (INIS)

    Full text: XRF Control and calibration software has come a long way in recent years. Advances in the multiple regression software sophistication and speed of computers have provided an essential resource to the XRF analyst. Over recent years there has been a trend amongst some analysts to develop XRF calibrations based exclusively on the statistical information given by calibration software. Multiple regression statistics are designed for non-correlated data sets but give unpredictable results if there are significant correlations in the standards used. This is typical for calibrations weighted towards certified reference materials (CRM's) which, being natural materials, contain correlated concentrations. Purely statistical methods in calibration development have applicability only over very short concentration ranges and for materials whose composition varies little. Beyond these ranges, the calibration has the potential to be unstable and has been known to produce significant deviations in analysis of unknown samples. The statistical information generated during XRF calibrations can be a very useful tool when used in conjunction with knowledge of the physics behind the correction factors applied. The matrix coefficients represent physical absorption/enhancement effects within the sample and are not arbitrary numbers used to get a good fit to the calibration line. Inappropriate use of matrix factors and overlap factors can produce low RMS values but erroneous results in unknown samples. This talk will contain examples to demonstrate hazards with different calibration strategies and will include coverage of the following topics: physical effects occurring within the sample as a result of X-ray irradiation; use of multiple regression statistics and what role it plays in the calibration; calibration strategies using synthetic and CRM standards; determining appropriate theoretical and semi-empirical matrix corrections and line overlap factors. Copyright (1999

  13. SPRT Calibration Uncertainties and Internal Quality Control at a Commercial SPRT Calibration Facility

    Science.gov (United States)

    Wiandt, T. J.

    2008-06-01

    The Hart Scientific Division of the Fluke Corporation operates two accredited standard platinum resistance thermometer (SPRT) calibration facilities, one at the Hart Scientific factory in Utah, USA, and the other at a service facility in Norwich, UK. The US facility is accredited through National Voluntary Laboratory Accreditation Program (NVLAP), and the UK facility is accredited through UKAS. Both provide SPRT calibrations using similar equipment and procedures, and at similar levels of uncertainty. These uncertainties are among the lowest available commercially. To achieve and maintain low uncertainties, it is required that the calibration procedures be thorough and optimized. However, to minimize customer downtime, it is also important that the instruments be calibrated in a timely manner and returned to the customer. Consequently, subjecting the instrument to repeated calibrations or extensive repeated measurements is not a viable approach. Additionally, these laboratories provide SPRT calibration services involving a wide variety of SPRT designs. These designs behave differently, yet predictably, when subjected to calibration measurements. To this end, an evaluation strategy involving both statistical process control and internal consistency measures is utilized to provide confidence in both the instrument calibration and the calibration process. This article describes the calibration facilities, procedure, uncertainty analysis, and internal quality assurance measures employed in the calibration of SPRTs. Data will be reviewed and generalities will be presented. Finally, challenges and considerations for future improvements will be discussed.

  14. Calibrating pen dosimeters with and without a phantom

    Energy Technology Data Exchange (ETDEWEB)

    Nonato, Fernanda B.C.; Cescon, Claudinei T.; Caldas, Linda V.E., E-mail: fbnonato@ipen.b, E-mail: ctcescon@ipen.b, E-mail: lcaldas@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2011-07-01

    Thirty one direct reading dosimeters (pen dosimeters) were calibrated and tested in standard beams of gamma radiation, with and without the use of a phantom. The calibration was performed with a Co-60 source and tested with a Cs-137 source. The dose-response curves of the pen dosimeters and their calibration factors for a Co-60 source, with and without the use of a phantom were obtained. The results show the need to calibrate the pen dosimeters with a phantom. (author)

  15. Colour calibration for colour reproduction

    OpenAIRE

    Emmel, P.; R. D. Hersch

    2000-01-01

    Due to the proliferation of low-cost colour devices (digital colour cameras, scanners, printers etc.) during the last few years, colour calibration has become an important issue. Such devices should faithfully reproduce colour images, but experience shows they don't. Among the main reasons, we note the diversity of acquisition, display and printing technologies which makes standardization difficult. Each device has a different gamut, i.e. a different set of colours that it can acquire or repr...

  16. Spiral reader calibration

    International Nuclear Information System (INIS)

    The method to calibrate the spiral reader (SR) is presented. A brief description of the main procedures of the calibration program SCALP, adapted for the IHEP equipment and purposes, is described. The precision characteristics of the IHEP SR have been analysed on the results, presented in the form of diagrams. There is a calibration manual for the user

  17. Cumulative sum quality control for calibrated breast density measurements

    Energy Technology Data Exchange (ETDEWEB)

    Heine, John J.; Cao Ke; Beam, Craig [Cancer Prevention and Control Division, Moffitt Cancer Center, 12902 Magnolia Drive, Tampa, Florida 33612 (United States); Division of Epidemiology and Biostatistics, School of Public Health, University of Illinois at Chicago, 1603 W. Taylor St., Chicago, Illinois 60612 (United States)

    2009-12-15

    Purpose: Breast density is a significant breast cancer risk factor. Although various methods are used to estimate breast density, there is no standard measurement for this important factor. The authors are developing a breast density standardization method for use in full field digital mammography (FFDM). The approach calibrates for interpatient acquisition technique differences. The calibration produces a normalized breast density pixel value scale. The method relies on first generating a baseline (BL) calibration dataset, which required extensive phantom imaging. Standardizing prospective mammograms with calibration data generated in the past could introduce unanticipated error in the standardized output if the calibration dataset is no longer valid. Methods: Sample points from the BL calibration dataset were imaged approximately biweekly over an extended timeframe. These serial samples were used to evaluate the BL dataset reproducibility and quantify the serial calibration accuracy. The cumulative sum (Cusum) quality control method was used to evaluate the serial sampling. Results: There is considerable drift in the serial sample points from the BL calibration dataset that is x-ray beam dependent. Systematic deviation from the BL dataset caused significant calibration errors. This system drift was not captured with routine system quality control measures. Cusum analysis indicated that the drift is a sign of system wear and eventual x-ray tube failure. Conclusions: The BL calibration dataset must be monitored and periodically updated, when necessary, to account for sustained system variations to maintain the calibration accuracy.

  18. Calibration of Nanopositioning Stages

    Directory of Open Access Journals (Sweden)

    Ning Tan

    2015-12-01

    Full Text Available Accuracy is one of the most important criteria for the performance evaluation of micro- and nanorobots or systems. Nanopositioning stages are used to achieve the high positioning resolution and accuracy for a wide and growing scope of applications. However, their positioning accuracy and repeatability are not well known and difficult to guarantee, which induces many drawbacks for many applications. For example, in the mechanical characterisation of biological samples, it is difficult to perform several cycles in a repeatable way so as not to induce negative influences on the study. It also prevents one from controlling accurately a tool with respect to a sample without adding additional sensors for closed loop control. This paper aims at quantifying the positioning repeatability and accuracy based on the ISO 9283:1998 standard, and analyzing factors influencing positioning accuracy onto a case study of 1-DoF (Degree-of-Freedom nanopositioning stage. The influence of thermal drift is notably quantified. Performances improvement of the nanopositioning stage are then investigated through robot calibration (i.e., open-loop approach. Two models (static and adaptive models are proposed to compensate for both geometric errors and thermal drift. Validation experiments are conducted over a long period (several days showing that the accuracy of the stage is improved from typical micrometer range to 400 nm using the static model and even down to 100 nm using the adaptive model. In addition, we extend the 1-DoF calibration to multi-DoF with a case study of a 2-DoF nanopositioning robot. Results demonstrate that the model efficiently improved the 2D accuracy from 1400 nm to 200 nm.

  19. Electromagnetic Calorimeter Calibration with $\\pi^{0}$

    CERN Multimedia

    Puig Navarro, A

    2009-01-01

    Several methods can be used in order to achieve precise calibration of the LHCb Electromagnetic Calorimeter (ECAL) once reasonable cell equalization has been reached. At low transverse energy, the standard calibration procedure is an iterative method based on the fit of the $\\gamma\\gamma$ invariant mass distribution for each cell of the decay $\\pi^{0}\\to\\gamma\\gamma$ with resolved photons. A new technique for generating the combinatorial background of such decays directly from data has been developed. Knowledge of the background could allow an alternative calibration method based on a event by event fit of the same $\\gamma\\gamma$ invariant mass distribution where contributions from groups of cells are considered in a single fit. The background generation procedure and this possible new calibration method are presented in this poster, in addition to an overview of the LHCb Calorimetry system and ECAL calibration techniques.

  20. Calibration of Sr-90 ophthalmic applicators

    International Nuclear Information System (INIS)

    The purpose of this paper is to alert users of Sr-90 ophthalmic applicators about potential large errors in calibration of these devices. A discrepancy of more than 50% in calibration of Sr-90 ophthalmic applicators between the US National Institute of Standards and Technology (NIST) and one foreign manufacturer (the world's only remaining supplier) has been reported. A single-plane Sr-90 ophthalmic applicator was calibrated by the manufacturer, by NIST, and by the University of Wisconsin. The manufacturer's close rate calibration is confined to a 3-mm-diameter active area, while NIST measures all beta radiation emitted into a 2-PI solid angle. The discrepancy was verified by means of a technique based on that of NIST. Reports of calibrations at NIST of applicators made by several American manufacturers (no longer available) indicate that large discrepancies exist for other manufacturers as well

  1. Residual gas analyzer calibration

    Science.gov (United States)

    Lilienkamp, R. H.

    1972-01-01

    A technique which employs known gas mixtures to calibrate the residual gas analyzer (RGA) is described. The mass spectra from the RGA are recorded for each gas mixture. This mass spectra data and the mixture composition data each form a matrix. From the two matrices the calibration matrix may be computed. The matrix mathematics requires the number of calibration gas mixtures be equal to or greater than the number of gases included in the calibration. This technique was evaluated using a mathematical model of an RGA to generate the mass spectra. This model included shot noise errors in the mass spectra. Errors in the gas concentrations were also included in the valuation. The effects of these errors was studied by varying their magnitudes and comparing the resulting calibrations. Several methods of evaluating an actual calibration are presented. The effects of the number of gases in then, the composition of the calibration mixture, and the number of mixtures used are discussed.

  2. Selection of stars to calibrate Gaia

    Science.gov (United States)

    Carrasco, J. M.; Voss, H.; Jordi, C.; Fabricius, C.; Pancino, E.; Altavilla, G.

    2015-05-01

    Gaia is an all-sky survey satellite, launched by ESA on 19th December 2013, to obtain parallaxes and proper motions to microarcsecond level precision, radial velocities and astrophysical parameters for about one billion objects down to a limiting magnitude of 20. The chosen strategy to perform the photometric calibration is to split the process into two steps, internal and external calibration. The internal calibration will combine all different transits of a given source to a common reference internal system producing a 'mean' Gaia observation. This internal calibration accounts for the differential instrumental effects (in sensitivity, aperture, PSF, etc.). They depend on the colour and type of the source. For this reason, a selection of calibration sources ensuring a good representation of all kind of observed sources is needed. The entire magnitude and colour range of the sources have to be covered by these calibration stars and for all calibration intervals. It is a challenge to obtain a suitable colour distribution for the standards, especially for bright sources and the daily large scale calibration intervals. Once the mean Gaia observations are produced, a final step, the external calibration, transforms them to absolute fluxes and wavelengths. In principle, few calibration sources are needed (about 200 spectrophotometric standard stars, SPSS, are currently being considered). They need to have accurate determinations of their absolute fluxes and their non-variability need to be ensured below 1% precision. For this purpose, a big international observational effort is being done (using telescopes as 2.2m@CAHA, TNG@LaPalma, NTT@LaSilla, LaRuca@SPM, and others). During this observational effort some cases of non-expected variability of the SPSS candidates have been discovered.

  3. HPS instrument calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Masse, F.X; Eisenhower, E.H.; Swinth, K.L.

    1993-12-31

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory.

  4. HPS instrument calibration laboratory accreditation program

    International Nuclear Information System (INIS)

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory

  5. Preliminary evaluation of a Neutron Calibration Laboratory

    Energy Technology Data Exchange (ETDEWEB)

    Alvarenga, Talysson S.; Neves, Lucio P.; Perini, Ana P.; Sanches, Matias P.; Mitake, Malvina B.; Caldas, Linda V.E., E-mail: talvarenga@ipen.br, E-mail: lpneves@ipen.br, E-mail: aperini@ipen.br, E-mail: msanches@ipen.br, E-mail: mbmitake@ipen.br, E-mail: lcaldas@ipen.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Federico, Claudio A., E-mail: claudiofederico@ieav.cta.br [Instituto de Estudos Avancados (IEAv/DCTA), Sao Jose dos Campos, SP (Brazil). Dept. de Ciencia e Tecnologia Aeroespacial

    2013-07-01

    In the past few years, Brazil and several other countries in Latin America have experimented a great demand for the calibration of neutron detectors, mainly due to the increase in oil prospection and extraction. The only laboratory for calibration of neutron detectors in Brazil is localized at the Institute for Radioprotection and Dosimetry (IRD/CNEN), Rio de Janeiro, which is part of the IAEA SSDL network. This laboratory is the national standard laboratory in Brazil. With the increase in the demand for the calibration of neutron detectors, there is a need for another calibration services. In this context, the Calibration Laboratory of IPEN/CNEN, Sao Paulo, which already offers calibration services of radiation detectors with standard X, gamma, beta and alpha beams, has recently projected a new calibration laboratory for neutron detectors. In this work, the ambient equivalent dose rate (H⁎(10)) was evaluated in several positions inside and around this laboratory, using Monte Carlo simulation (MCNP5 code), in order to verify the adequateness of the shielding. The obtained results showed that the shielding is effective, and that this is a low-cost methodology to improve the safety of the workers and evaluate the total staff workload. (author)

  6. Calibration of Ground-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...... measurements are given for information only....

  7. Calibration of Ground-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  8. AXAF calibration: the HXDS flow proportional counters

    Science.gov (United States)

    Wargelin, Bradford J.; Kellogg, Edwin M.; McDermott, Walter C.; Evans, Ian N.; Vitek, S. A.

    1997-07-01

    The design, performance, and calibration of the seven flow proportional counters (FPCs) used during AXAF ground calibration are described. Five of the FPCs served as beam normalization detectors (BNDs), and two were used in the telescope focal plane in combination with a set of apertures to measure the point response functions and effective areas of the AXAF mirrors and transmission gratings. The BNDs also provide standards for determining the effective areas of the several telescope/grating/flight-detector combinations. With useful energy resolution and quantum efficiency over the entire 100-eV to 10 keV AXAF energy band, the FPCs provided most of the data acquired during AXAF calibration. Although the principles of proportional counter operation are relatively simple, AXAF's stringent calibration goals require detailed calibration and modeling of such effects as window- support-wire obscuration, window deformation between the support wires, electron diffusion and avalanche processes, gain nonuniformities, and gas pressure and temperature variations. Detector aperture areas and signal processing deadtime must also be precisely determined, and detector degradation during the many months of AXAF calibration must be prevented. The FPC calibration program is based on measurement of individual components (such as window transmission and aperture size) and the relative quantum efficiencies of complete detector systems, as well as absolute QE calibration of selected detectors at the BESSY synchrotron, an x-ray source of precisely known intensity.

  9. Wind Tunnel Force Balance Calibration Study - Interim Results

    Science.gov (United States)

    Rhew, Ray D.

    2012-01-01

    Wind tunnel force balance calibration is preformed utilizing a variety of different methods and does not have a direct traceable standard such as standards used for most calibration practices (weights, and voltmeters). These different calibration methods and practices include, but are not limited to, the loading schedule, the load application hardware, manual and automatic systems, re-leveling and non-re-leveling. A study of the balance calibration techniques used by NASA was undertaken to develop metrics for reviewing and comparing results using sample calibrations. The study also includes balances of different designs, single and multi-piece. The calibration systems include, the manual, and the automatic that are provided by NASA and its vendors. The results to date will be presented along with the techniques for comparing the results. In addition, future planned calibrations and investigations based on the results will be provided.

  10. Calibration of two 90Sr+90Y dermatological applicators

    International Nuclear Information System (INIS)

    The 90Sr+90Y applicators need to be periodically calibrated, but in Brazil the service it not offered yet. The recommended method for the calibration of this kind of applicators is the use of extrapolation chambers. An alternative method for the calibration of clinical applicators is the use of thermoluminescent dosimeters. A dosimetric method of these applicators was already developed at Instituto de Pesquisas Energeticas e Nucleares (IPEN) and several types of thermoluminescent dosimeters were studied in previous works. The aim of this work was the application of this method to calibrate two dermatological applicators. Thin CaSO4:Dy pellets, with and without 10% of graphite were utilized. The reproducibility of these pellets was studied, and calibration curves were obtained using a standard applicator calibrated at the National Institute of Standards and Technology (NIST), USA. Both applicators showed similar results. The TL materials tested showed usefulness for dosimetry and calibration of this kind of applicators. (author)

  11. PACS photometer calibration block analysis

    CERN Document Server

    Moór, A; Kiss, Cs; Balog, Z; Billot, N; Marton, G

    2013-01-01

    The absolute stability of the PACS bolometer response over the entire mission lifetime without applying any corrections is about 0.5% (standard deviation) or about 8% peak-to-peak. This fantastic stability allows us to calibrate all scientific measurements by a fixed and time-independent response file, without using any information from the PACS internal calibration sources. However, the analysis of calibration block observations revealed clear correlations of the internal source signals with the evaporator temperature and a signal drift during the first half hour after the cooler recycling. These effects are small, but can be seen in repeated measurements of standard stars. From our analysis we established corrections for both effects which push the stability of the PACS bolometer response to about 0.2% (stdev) or 2% in the blue, 3% in the green and 5% in the red channel (peak-to-peak). After both corrections we still see a correlation of the signals with PACS FPU temperatures, possibly caused by parasitic h...

  12. Use of Ethanol/Diesel Blend and Advanced Calibration Methods to Satisfy Euro 5 Emission Standards without DPF Utilisation d’un carburant Diesel éthanolé à l’aide de méthodes de calibration avancées afin de satisfaire les normes Euro 5 sans filtre à particules

    Directory of Open Access Journals (Sweden)

    Magand S.

    2011-11-01

    innovative calibration methods, based on the simultaneous optimisation of engine basic settings and cold correction maps, are introduced in order to better suit to the new formulation impact on combustion and catalyst light-off and to drop off engine-out unburned hydrocarbons and carbon monoxide emissions. This stage allows pushing forward the work on test bed facilities in order to reduce the amount of vehicle tests. Tests on a chassis dynamometer are only used to validate the engine test bed results and to perform final tuning of cold correction maps. This alternative blend shows potential to achieve Euro 5 standard with Euro 4 Diesel vehicle configuration, without any hardware modification and without a Diesel particulate filter in the exhaust line. Such an innovative fuel formulation seems to be an interesting answer to the trade-off in the forthcoming years between cost and emissions reduction to achieve sustainable mobility. The presented calibration methods and tools allow to fully take advantage of this alternative fuel in a reduced time scale. L’utilisation des biocarburants s’est développée durant ces dernières années de façon importante afin de diversifier les sources d’énergies et de limiter la hausse des émissions de gaz à effet de serre du secteur des transports. L’un des carburants renouvelables les plus adaptés à une production de masse est l’éthanol. Celui-ci est aujourd’hui principalement utilisé dans les moteurs à allumage commandé, alors que la part des véhicules Diesel sur le marché européen est de l’ordre de 60 %. Ce constat nous a incité à proposer une formulation innovante utilisant de l’éthanol pour les applications Diesel. Les principaux verrous technologiques pour cette utilisation sont la miscibilité, la température d’éclair, la lubrification ou encore l’indice de cétane. Des travaux ont été réalisés pour optimiser la formulation contenant de l’éthanol, des biodiesels de première et seconde g

  13. European intercomparison of diagnostic dosemeters: calibration of the reference dosemeters

    International Nuclear Information System (INIS)

    The paper describes both the organisational and technical steps taken in the calibration of the reference dosemeters used in the intercomparison of diagnostic dosemeters carried out in Europe in 1990. A brief account is given on the activities by the coordinating centres in the 19 countries and by the calibrating institution to establish the operational framework necessary for carrying out the calibration. The second part describes the calibration of the dosemeters against PTB's primary standards in terms of the equipment used and of the individual steps of measurement taken. An assessment of the uncertainties associated with the calibration is presented

  14. Calibration of radiation protection monitoring equipment

    International Nuclear Information System (INIS)

    The facilities and equipment required for calibrating radiation protection monitoring equipment are both complex and expensive. The staff required to operate the equipment and perform the standardization have to be experienced, well trained and technically competent in the use of secondary or primary radiation standards and equipment. Thus an organization cannot commit considerable resources to perform calibration work purely for academic or scientific reasons. In most cases it must be justified by legal requirements pertaining within the country. These legal requirements on measuring devices for radiation protection purposes may be broadly divided into those relating to instruments intended for area monitoring and to those intended for individual monitoring

  15. OLI Radiometric Calibration

    Science.gov (United States)

    Markham, Brian; Morfitt, Ron; Kvaran, Geir; Biggar, Stuart; Leisso, Nathan; Czapla-Myers, Jeff

    2011-01-01

    Goals: (1) Present an overview of the pre-launch radiance, reflectance & uniformity calibration of the Operational Land Imager (OLI) (1a) Transfer to orbit/heliostat (1b) Linearity (2) Discuss on-orbit plans for radiance, reflectance and uniformity calibration of the OLI

  16. ACCESS: Design, Calibration Strategy, and Status

    Science.gov (United States)

    Kaiser, M. E.; Access Team

    2016-05-01

    ACCESS, Absolute Color Calibration Experiment for Standard Stars, is a series of rocket-borne sub-orbital missions and ground-based experiments designed to enable improvements in the precision of the astrophysical flux scale through the transfer of absolute laboratory detector standards from the National Institute of Standards and Technology (NIST) to a network of stellar standards with a calibration accuracy of 1% and a spectral resolving power of 500 across the 0.35–1.7μm bandpass. Establishing improved spectrophotometric standards is important for a broad range of missions and is relevant to many astrophysical problems. Systematic errors associated with problems such as dark energy now compete with the statistical errors and thus limit our ability to answer fundamental questions in astrophysics.

  17. The GERDA calibration system

    International Nuclear Information System (INIS)

    A system with three identical custom made units is used for the energy calibration of the GERDA Ge diodes. To perform a calibration the 228Th sources are lowered from the parking positions at the top of the cryostat. Their positions are measured by two independent modules. One, the incremental encoder, counts the holes in the perforated steel band holding the sources, the other measures the drive shaft's angular position even if not powered. The system can be controlled remotely by a Labview program. The calibration data is analyzed by an iterative calibration algorithm determining the calibration functions for different energy reconstruction algorithms and the resolution of several peaks in the 228Th spectrum is determined. A Monte Carlo simulation using the GERDA simulation software MAGE has been performed to determine the background induced by the sources in the parking positions.

  18. The PREMOS/PICARD instrument calibration

    Science.gov (United States)

    Schmutz, Werner; Fehlmann, André; Hülsen, Gregor; Meindl, Peter; Winkler, Rainer; Thuillier, Gérard; Blattner, Peter; Buisson, François; Egorova, Tatiana; Finsterle, Wolfgang; Fox, Nigel; Gröbner, Julian; Hochedez, Jean-François; Koller, Silvio; Meftah, Mustapha; Meisonnier, Mireille; Nyeki, Stephan; Pfiffner, Daniel; Roth, Hansjörg; Rozanov, Eugene; Spescha, Marcel; Wehrli, Christoph; Werner, Lutz; Wyss, Jules U.

    2009-08-01

    PREMOS is a space experiment scheduled to fly on the French solar mission PICARD. The experiment comprises filter radiometers and absolute radiometers to measure the spectral and total solar irradiance. The aim of PREMOS is to contribute to the long term monitoring of the total solar irradiance, to use irradiance observations for 'nowcasting' the state of the terrestrial middle atmosphere and to provide long term sensitivity calibration for the solar imaging instrument SODISM on PICARD. In this paper we describe the calibration of the instruments. The filter radiometer channels in the visible and near IR were characterized at PMOD/WRC and the UV channels were calibrated at PTB Berlin. The absolute radiometers were compared with the World Radiometric Reference at PMOD/WRC and a power calibration relative to a primary cryogenic radiometer standard was performed in vacuum and air at NPL.

  19. Protocols for calibrating multibeam sonar

    OpenAIRE

    Foote, Kenneth G.; Chu, Dezhang; Hammar, Terence R.; Baldwin, Kenneth C.; Mayer, Larry A.; Hufnagle, Lawrence C. jr.; Jech, J. Michael

    2005-01-01

    Development of protocols for calibrating multibeam sonar by means of the standard-target method is documented. Particular systems used in the development work included three that provide the water-column signals, namely the SIMRAD SM2000/90- and 200-kHz sonars and RESON SeaBat 8101 sonar, with operating frequency of 240 kHz. Two facilities were instrumented specifically for the work: a sea well at the Woods Hole Oceanographic Institution and a large, indoor freshwater tank at the University o...

  20. Segment Based Camera Calibration

    Institute of Scientific and Technical Information of China (English)

    马颂德; 魏国庆; 等

    1993-01-01

    The basic idea of calibrating a camera system in previous approaches is to determine camera parmeters by using a set of known 3D points as calibration reference.In this paper,we present a method of camera calibration in whih camera parameters are determined by a set of 3D lines.A set of constraints is derived on camea parameters in terms of perspective line mapping.Form these constraints,the same perspective transformation matrix as that for point mapping can be computed linearly.The minimum number of calibration lines is 6.This result generalizes that of Liu,Huang and Faugeras[12] for camera location determination in which at least 8 line correspondences are required for linear computation of camera location.Since line segments in an image can be located easily and more accurately than points,the use of lines as calibration reference tends to ease the computation in inage preprocessing and to improve calibration accuracy.Experimental results on the calibration along with stereo reconstruction are reported.

  1. Photometric Calibration of the Swift Ultraviolet/Optical Telescope

    CERN Document Server

    Poole, T S; Page, M J; Landsman, W; Holland, S T; Roming, P; Kuin, N P M; Brown, P J; Gronwall, C; Hunsberger, S; Koch, S; Mason, K O; Schady, P; Berk, D Vanden; Blustin, A J; Boyd, P; Broos, P; Carter, M; Chester, M M; Cucchiara, A; Hancock, B; Huckle, H; Immler, S; Ivanushkina, M; Kennedy, T; Marshall, F; Morgan, A; Pandey, S; de Pasquale, M; Smith, P J; Still, M

    2007-01-01

    We present the photometric calibration of the Swift UltraViolet/Optical Telescope (UVOT) which includes: optimum photometric and background apertures, effective area curves, colour transformations, conversion factors for count rates to flux, and the photometric zero points (which are accurate to better than 4 per cent) for each of the seven UVOT broadband filters. The calibration was performed with observations of standard stars and standard star fields that represent a wide range of spectral star types. The calibration results include the position dependent uniformity, and instrument response over the 1600-8000A operational range. Because the UVOT is a photon counting instrument, we also discuss the effect of coincidence loss on the calibration results. We provide practical guidelines for using the calibration in UVOT data analysis. The results presented here supersede previous calibration results.

  2. Calibration of dermatological applicators of 90 Sr+90 Y

    International Nuclear Information System (INIS)

    90 Sr+90 Y dermatological applicators are widely used in the treatment of skin lesions. Despite calibrated by the manufacturers, these sources must be re-calibrated periodically by standard laboratories. Articles published by different authors show the discrepancies between manufacturers and standard laboratories calibrations of 90 Sr+90 Y applicators. Ionization chambers with variable volume, named extrapolation chambers, are utilized for the calibration of such sources. An extrapolation chamber was developed at IPEN for the calibration of 90 Sr+90 Y dermatological applicators. This chamber shows a good performance in the detection of beta particles. The aim of this work is to establish and to apply a routine calibration procedure to a dermatological applicator, based on former work developed in this institution. (author)

  3. A Novel Calibrator for Electronic Transformers Based on IEC 61850

    Directory of Open Access Journals (Sweden)

    Baoxiang PAN

    2013-01-01

    Full Text Available It is necessary for electronic transformer to make calibration before putting it into practice. To solve the problems in actual calibration process, a novel electronic transformer calibrator is designed. In principle, this system adopts both the direct method and the difference method, which are two popular methods for electronic transformer calibration, by this way the application of the system is extended with its reliability improved. In the system design, based on virtual instrument technology, LabVIEW and WinPCap toolkit are used to develop the application software, and it is able to calibrate those electronic transformers following the standard of IEC 61850. In the calculation of ratio and phase error based on fast Fourier transform, a new window function is introduced, and thus the accuracy of calibration, influenced by the frequency vibration, is improved. This research provides theoretic support and practical reference to the development of intelligent calibrator for electronic transformers.

  4. An automated linearity test for direct voltage calibrators

    Science.gov (United States)

    Endsley, Ross

    1990-04-01

    The complete calibration of direct voltage calibrators should include a linearity calibration. The method described uses a string of thermally lagged resistors in series across a stable voltage source to provide fixed reference points for the calibration. A high-resolution digital multimeter is used to compare voltage increments in the output of the calibrator. The calibrator and the instruments in the linearity test system are controlled by computer via the IEEE-488 bus. The measured linearity is the result of standard ration techniques. The test is suggested not only as a general purpose technique, but also as a source of assistance in dealing with traceability questions arising over instruments employing international calibration routines.

  5. Lidar Calibration Centre

    Science.gov (United States)

    Pappalardo, Gelsomina; Freudenthaler, Volker; Nicolae, Doina; Mona, Lucia; Belegante, Livio; D'Amico, Giuseppe

    2016-06-01

    This paper presents the newly established Lidar Calibration Centre, a distributed infrastructure in Europe, whose goal is to offer services for complete characterization and calibration of lidars and ceilometers. Mobile reference lidars, laboratories for testing and characterization of optics and electronics, facilities for inspection and debugging of instruments, as well as for training in good practices are open to users from the scientific community, operational services and private sector. The Lidar Calibration Centre offers support for trans-national access through the EC HORIZON2020 project ACTRIS-2.

  6. Equipment for dosemeter calibration

    International Nuclear Information System (INIS)

    The device is used for precise calibration of dosimetric instrumentation, such as used at nuclear facilities. The high precision of the calibration procedure is primarily due to the fact that one single and steady radiation source is used. The accurate alignment of the source and the absence of shielding materials in the beam axis make for high homogeneity of the beam and reproducibility of the measurement; this is also contributed to by the horizontal displacement of the optical bench, which ensures a constant temperature field and the possibility of adjusting the radiation source at a sufficient distance from the instrument to be calibrated. (Z.S.). 3 figs

  7. In-Field Absolute Calibration of Ground and Airborne VIS-NIR-SWIR Hyperspectral Point Spectrometers

    Directory of Open Access Journals (Sweden)

    Offer Rozenstein

    2014-01-01

    Full Text Available Spectrometer calibration and measurements of spectral radiance are often required when performing ground, aerial, and space measurements. While calibrating a spectrometer in the field using an integrating sphere is practically unachievable, calibration against a quartz halogen (QH lamp is a quite easy and feasible option. We describe a calibration protocol whereby a professional QH lamp, operating with a stabilized current source, is first calibrated in the laboratory against a US National Institute of Standards and Technology (NIST traceable integrating sphere and, then, used for the field calibration of a spectrometer before a ground or airborne campaign. Another advantage of the lamp over the integrating sphere is its ability to create a continuous calibration curve at the spectrometer resolution, while the integrating sphere is calibrated only for a few discrete wavelengths. A calibrated lamp could also be used for a secondary continuous calibration of an un-calibrated integrating sphere.

  8. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina; Webb, Craig

    2016-05-02

    This presentation provides a high-level overview of the progress on the Broadband Outdoor Radiometer Calibrations for all shortwave and longwave radiometers that are deployed by the Atmospheric Radiation Measurement program.

  9. Air Data Calibration Facility

    Data.gov (United States)

    Federal Laboratory Consortium — This facility is for low altitude subsonic altimeter system calibrations of air vehicles. Mission is a direct support of the AFFTC mission. Postflight data merge is...

  10. Calibration of sound velocimeter in pure water

    Science.gov (United States)

    Li, Zhiwei; Zhang, Baofeng; Li, Tao; Zhu, Junchao; Xie, Ziming

    2016-01-01

    Accurate measurement of sound speed is important to calibrate a sound velocity profiler which provides real-time sound velocity to the sonar equipment in oceanographic survey. The sound velocity profiler calculates the sound speed by measuring the time-of-flight of a 1 MHz single acoustic pulse to travel over about 300 mm path. A standard sound velocimeter instrument was invited to calibrate the sound velocity profiler in pure water at temperatures of 278,283, 288, 293, 298, 303 and 308K in a thermostatic vessel at one atmosphere. The sound velocity profiler was deployed in the thermostatic vessel alongside the standard sound velocimeter instrument and two platinum resistance thermometers (PRT) which were calibrated to 0.002k by comparison with a standard PRT. Time of flight circuit board was used to measure the time-of-flight to 22 picosecond precision. The sound speed which was measured by the sound velocity profiler was compared to the standard sound speed calculated by UNESCO to give the laboratory calibration coefficients and was demonstrated agreement with CTD-derived sound speed using Del Grosso's seawater equation after removing a bias.

  11. Essay on Option Pricing, Hedging and Calibration

    DEFF Research Database (Denmark)

    da Silva Ribeiro, André Manuel

    -functional Interest Rate Model with Fast Fourier Transform This paper shows how the fast Fourier Transform (FFT) may be used to calibrate an interest rate Markov-functional (MF) model when the characteristic function of the Markov process is analytically known. We added a compound Poisson process to the standard...

  12. 14 CFR 33.85 - Calibration tests.

    Science.gov (United States)

    2010-01-01

    ... Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF TRANSPORTATION AIRCRAFT AIRWORTHINESS STANDARDS: AIRCRAFT ENGINES Block Tests; Turbine Aircraft Engines § 33.85 Calibration tests. (a) Each engine... tests form the basis for establishing the characteristics of the engine over its entire operating...

  13. Probability based calibration of pressure coefficients

    DEFF Research Database (Denmark)

    Hansen, Svend Ole; Pedersen, Marie Louise; Sørensen, John Dalsgaard

    2015-01-01

    not depend on the type of variable action. A probability based calibration of pressure coefficients have been carried out using pressure measurements on the standard CAARC building modelled on scale of 1:383. The extreme pressures measured on the CAARC building model in the wind tunnel have been fitted...

  14. Approximation Behooves Calibration

    DEFF Research Database (Denmark)

    da Silva Ribeiro, André Manuel; Poulsen, Rolf

    2013-01-01

    Calibration based on an expansion approximation for option prices in the Heston stochastic volatility model gives stable, accurate, and fast results for S&P500-index option data over the period 2005–2009.......Calibration based on an expansion approximation for option prices in the Heston stochastic volatility model gives stable, accurate, and fast results for S&P500-index option data over the period 2005–2009....

  15. Scanner calibration revisited

    Directory of Open Access Journals (Sweden)

    Pozhitkov Alexander E

    2010-07-01

    Full Text Available Abstract Background Calibration of a microarray scanner is critical for accurate interpretation of microarray results. Shi et al. (BMC Bioinformatics, 2005, 6, Art. No. S11 Suppl. 2. reported usage of a Full Moon BioSystems slide for calibration. Inspired by the Shi et al. work, we have calibrated microarray scanners in our previous research. We were puzzled however, that most of the signal intensities from a biological sample fell below the sensitivity threshold level determined by the calibration slide. This conundrum led us to re-investigate the quality of calibration provided by the Full Moon BioSystems slide as well as the accuracy of the analysis performed by Shi et al. Methods Signal intensities were recorded on three different microarray scanners at various photomultiplier gain levels using the same calibration slide from Full Moon BioSystems. Data analysis was conducted on raw signal intensities without normalization or transformation of any kind. Weighted least-squares method was used to fit the data. Results We found that initial analysis performed by Shi et al. did not take into account autofluorescence of the Full Moon BioSystems slide, which led to a grossly distorted microarray scanner response. Our analysis revealed that a power-law function, which is explicitly accounting for the slide autofluorescence, perfectly described a relationship between signal intensities and fluorophore quantities. Conclusions Microarray scanners respond in a much less distorted fashion than was reported by Shi et al. Full Moon BioSystems calibration slides are inadequate for performing calibration. We recommend against using these slides.

  16. Energy calibration via correlation

    CERN Document Server

    Maier, Daniel

    2015-01-01

    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241 Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be le...

  17. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  18. A Method to Test Model Calibration Techniques

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, Ron; Polly, Ben; Neymark, Joel

    2016-08-26

    This paper describes a method for testing model calibration techniques. Calibration is commonly used in conjunction with energy retrofit audit models. An audit is conducted to gather information about the building needed to assemble an input file for a building energy modeling tool. A calibration technique is used to reconcile model predictions with utility data, and then the 'calibrated model' is used to predict energy savings from a variety of retrofit measures and combinations thereof. Current standards and guidelines such as BPI-2400 and ASHRAE-14 set criteria for 'goodness of fit' and assume that if the criteria are met, then the calibration technique is acceptable. While it is logical to use the actual performance data of the building to tune the model, it is not certain that a good fit will result in a model that better predicts post-retrofit energy savings. Therefore, the basic idea here is that the simulation program (intended for use with the calibration technique) is used to generate surrogate utility bill data and retrofit energy savings data against which the calibration technique can be tested. This provides three figures of merit for testing a calibration technique, 1) accuracy of the post-retrofit energy savings prediction, 2) closure on the 'true' input parameter values, and 3) goodness of fit to the utility bill data. The paper will also discuss the pros and cons of using this synthetic surrogate data approach versus trying to use real data sets of actual buildings.

  19. Critical issues for implementation of the standard NBR ISO/IEC 17025:2005 in Testing and Calibration Laboratory: case study at a public institution; Aspectos criticos para implantacao da norma NBR ISO/IEC 17025:2005 em laboratorio de ensaio e calibracao: estudo de caso em uma instituicao publica

    Energy Technology Data Exchange (ETDEWEB)

    Castro, Denise Confar Carvalho de

    2013-07-01

    The public institution aims to promote excellence in public management to contribute to the quality of services provided to its customers and to increase competitiveness in the country, as well as its international projection. A technical barrier to trade that can lead to dissatisfaction and achieve the reputation of the institution is failing the test or calibration results and measurement data, thereby accreditation is regarded as the first essential step to facilitate the mutual acceptance of test results and calibration or measurement data. For recognition, laboratories need to demonstrate full compliance with both the sections of ISO/IEC 17025:2005, i.e. management and technical requirements. This research aims to discuss the critical aspects for implementation of ABNT NBR ISO / IEC 17025:2005 for calibration and testing of a Public Institution seeking accreditation of its laboratories with INMETRO, national accreditation body Laboratories. Besides getting preventive, corrective and improvement actions continues guidelines. Furthermore, the methodology used was to conduct a literature search and apply a questionnaire to identify the degree of agreement / disagreement of the foundations of the standard servers. Analysis of the results showed that the critical issues were: commitment, training, resources (infrastructure, human) and culture. (author)

  20. Calibration of burnup monitor installed in Rokkasho Reprocessing Plant

    Energy Technology Data Exchange (ETDEWEB)

    Oeda, Kaoru; Naito, Hirofumi; Hirota, Masanari [Japan Nuclear Fuel Co. Ltd., Rokkasho, Aomori (Japan); Natsume, Koichiro [Isogo Engineering Center, Toshiba Corporation, Yokohama, Kanagawa (Japan); Kumanomido, Hironori [Nuclear Engineering Laboratory, Toshiba Corporation, Kawasaki, Kanagawa (Japan)

    2000-06-01

    Rokkasho Reprocessing Plant uses burnup credit for criticality control at the Spent Fuel Storage Facility (SFSF) and the Dissolution Facility. A burnup monitor measures nondestructively burnup value of a spent fuel assembly and guarantees the credit for burnup. For practical reasons, a standard radiation source is not used in calibration of the burnup monitor, but the burnup values of many spent fuel assemblies are measured based on operator-declared burnup values. This paper describes the concept of burnup credit, the burnup monitor, and the calibration method. It is concluded, from the results of calibration tests, that the calibration method is valid. (author)

  1. Calibration of ionization chambers used in LDR brachytherapy

    International Nuclear Information System (INIS)

    In this work was developed a calibration procedure of well-type ionization chambers used for measurements of I-125, seed type. It was used as a standard an ionization chamber Capintec CRC-15BT, with calibration certificate of the University of Wisconsin. It were calibrated two well-type ionization chambers of Capintec CRC-15R model. A source of I-125 was used in clinical use (18.5 to 7.4 MBq). The results showed that with the application of calibration factors was possible to decrease read deviate from 16% to just 1.0%

  2. Manual Calibration System for Daya Bay Reactor Neutrino Experiment

    Institute of Scientific and Technical Information of China (English)

    HUANG; Han-xiong; RUAN; Xi-chao; REN; Jie; LV; Yin-long; FAN; Cheng-jun; CHEN; Yan-nan; WANG; Zhao-hui; ZHOU; Zu-ying; HOU; Long; ZHANG; Jia-wen; ZHANG; Yin-hong; YU; Chao-ju; HE; Wei; ZHOU; Bin

    2012-01-01

    <正>The neutrino mixing angle θ13 with a significance of 7.7 standard deviations has been published by the Daya Bay anti-neutrino experiment collaboration in 2012. To understand the non-uniformity and the energy non-linearity of the anti-neutrino detector (AD), a calibration campaign for the AD1 with the Manual Calibration System (MCS) has been finished. The aim of this calibration plan is to deploy the calibration sources to any positions inside the Inner Acrylic Vessel (IAV), to study detail properties of AD.

  3. Measurement quality assurance for beta particle calibrations at NIST

    Energy Technology Data Exchange (ETDEWEB)

    Soares, C.G.; Pruitt, J.S. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    Standardized beta-particle fields have been established in an international standard and have been adopted for use in several U.S. dosimeter and instrument testing standards. Calibration methods and measurement quality assurance procedures employed at the National Institute of Standards and Technology (NIST) for beta-particle calibrations in these reference fields are discussed. The calibration facility including the NIST-automated extrapolation ionization chamber is described, and some sample results of calibrations are shown. Methods for establishing and maintaining traceability to NIST of secondary laboratories are discussed. Currently, there are problems in finding a good method for routine testing of traceability to NIST. Some examples of past testing methods are given and solutions to this problem are proposed.

  4. Photometric calibrations for 21st century science

    Energy Technology Data Exchange (ETDEWEB)

    Kent, Stephen; /Fermilab; Kaiser, Mary Elizabeth; /Johns Hopkins U.; Deustua, Susana E.; /Baltimore, Space Telescope Sci.; Smith, J.Allyn; /Austin Peay State U.; Adelman, Saul; /Citadel Military Coll.; Allam, Sahar S.; /Fermilab; Baptista, Brian; /Indiana U.; Bohlin, Ralph C.; /Baltimore, Space Telescope Sci.; Clem, James L.; /Louisiana State U.; Conley, Alex; /Colorado U.; Edelstein, Jerry; /UC, Berkeley, Space Sci. Dept. /NOAO, Tucson

    2009-02-01

    The answers to fundamental science questions in astrophysics, ranging from the history of the expansion of the universe to the sizes of nearby stars, hinge on our ability to make precise measurements of diverse astronomical objects. As our knowledge of the underlying physics of objects improves along with advances in detectors and instrumentation, the limits on our capability to extract science from measurements is set, not by our lack of understanding of the nature of these objects, but rather by the most mundane of all issues: the precision with which we can calibrate observations in physical units. In principle, photometric calibration is a solved problem - laboratory reference standards such as blackbody furnaces achieve precisions well in excess of those needed for astrophysics. In practice, however, transferring the calibration from these laboratory standards to astronomical objects of interest is far from trivial - the transfer must reach outside the atmosphere, extend over 4{pi} steradians of sky, cover a wide range of wavelengths, and span an enormous dynamic range in intensity. Virtually all spectrophotometric observations today are calibrated against one or more stellar reference sources, such as Vega, which are themselves tied back to laboratory standards in a variety of ways. This system's accuracy is not uniform. Selected regions of the electromagnetic spectrum are calibrated extremely well, but discontinuities of a few percent still exist, e.g., between the optical and infrared. Independently, model stellar atmospheres are used to calibrate the spectra of selected white dwarf stars, e.g. the HST system, but the ultimate accuracy of this system should be verified against laboratory sources. Our traditional standard star systems, while sufficient until now, need to be improved and extended in order to serve future astrophysics experiments. This white paper calls for a program to improve upon and expand the current networks of

  5. SI-Traceable Calibrations of Celestial Objects

    Science.gov (United States)

    Cramer, C. E.; Lykke, K. R.; Woodward, J. T.

    2016-05-01

    Photometric calibration is currently the leading source of systematic uncertainty in supernova surveys that aim to determine the nature of dark energy. The bulk of this uncertainty is due to imperfect knowledge of the spectral energy distribution of stars used as primary standards. We review the challenges associated with establishing an absolute calibration of stellar spectra and describe how it is possible to do better by using recent advances in optical metrology, paying particular attention to the measurement chain establishing SI-traceability and reporting of measurement uncertainties.

  6. HAWC Timing Calibration

    CERN Document Server

    Huentemeyer, Petra; Dingus, Brenda

    2009-01-01

    The High-Altitude Water Cherenkov (HAWC) Experiment is a second-generation highsensitivity gamma-ray and cosmic-ray detector that builds on the experience and technology of the Milagro observatory. Like Milagro, HAWC utilizes the water Cherenkov technique to measure extensive air showers. Instead of a pond filled with water (as in Milagro) an array of closely packed water tanks is used. The event direction will be reconstructed using the times when the PMTs in each tank are triggered. Therefore, the timing calibration will be crucial for reaching an angular resolution as low as 0.25 degrees.We propose to use a laser calibration system, patterned after the calibration system in Milagro. Like Milagro, the HAWC optical calibration system will use ~1 ns laser light pulses. Unlike Milagro, the PMTs are optically isolated and require their own optical fiber calibration. For HAWC the laser light pulses will be directed through a series of optical fan-outs and fibers to illuminate the PMTs in approximately one half o...

  7. Calibration Under Uncertainty.

    Energy Technology Data Exchange (ETDEWEB)

    Swiler, Laura Painton; Trucano, Timothy Guy

    2005-03-01

    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  8. Polarimetric Palsar Calibration

    Science.gov (United States)

    Touzi, R.; Shimada, M.

    2008-11-01

    Polarimetric PALSAR system parameters are assessed using data sets collected over various calibration sites. The data collected over the Amazonian forest permits validating the zero Faraday rotation hypotheses near the equator. The analysis of the Amazonian forest data and the response of the corner reflectors deployed during the PALSAR acquisitions lead to the conclusion that the antenna is highly isolated (better than -35 dB). Theses results are confirmed using data collected over the Sweden and Ottawa calibration sites. The 5-m height trihedrals deployed in the Sweden calibration site by the Chalmers University of technology permits accurate measurement of antenna parameters, and detection of 2-3 degree Faraday rotation during day acquisition, whereas no Faraday rotation was noted during night acquisition. Small Faraday rotation angles (2-3 degree) have been measured using acquisitions over the DLR Oberpfaffenhofen and the Ottawa calibration sites. The presence of small but still significant Faraday rotation (2-3 degree) induces a CR return at the cross-polarization HV and VH that should not be interpreted as the actual antenna cross-talk. PALSAR antenna is highly isolated (better than -35 dB), and diagonal antenna distortion matrices (with zero cross-talk terms) can be used for accurate calibration of PALSAR polarimetric data.

  9. GTC Photometric Calibration

    Science.gov (United States)

    di Cesare, M. A.; Hammersley, P. L.; Rodriguez Espinosa, J. M.

    2006-06-01

    We are currently developing the calibration programme for GTC using techniques similar to the ones use for the space telescope calibration (Hammersley et al. 1998, A&AS, 128, 207; Cohen et al. 1999, AJ, 117, 1864). We are planning to produce a catalogue with calibration stars which are suitable for a 10-m telescope. These sources will be not variable, non binary and do not have infrared excesses if they are to be used in the infrared. The GTC science instruments require photometric calibration between 0.35 and 2.5 microns. The instruments are: OSIRIS (Optical System for Imaging low Resolution Integrated Spectroscopy), ELMER and EMIR (Espectrógrafo Multiobjeto Infrarrojo) and the Acquisition and Guiding boxes (Di Césare, Hammersley, & Rodriguez Espinosa 2005, RevMexAA Ser. Conf., 24, 231). The catalogue will consist of 30 star fields distributed in all of North Hemisphere. We will use fields containing sources over the range 12 to 22 magnitude, and spanning a wide range of spectral types (A to M) for the visible and near infrared. In the poster we will show the method used for selecting these fields and we will present the analysis of the data on the first calibration fields observed.

  10. Calibration of the MACHO photometry database

    Energy Technology Data Exchange (ETDEWEB)

    Alcock, A

    1998-10-23

    The MACHO Project is a microlensing survey that monitors the brightnesses of ~60 million stars in the Large Magellanic Cloud (LMC), Small Magellanic Cloud, and Galactic bulge. The database presently contains more photometric measurements than previously recorded in the history of astronomy. We describe the calibration of the MACHO two-color photometry and transformation to the standard Kron-Cousins V and R system. This allows for proper comparison with all other observations on the Kron-Cousins standard system. The highest precision calibrations are for ~9 million stars in the LMC bar. For these stars, independent photometric measurements in field-overlap regions indicate standard deviations δvR = 0.020 mag. Calibrated MACHO photometry data are compared with published photometric sequences and with new Hubble Space Telescope observations. We additionally describe the first application of these calibrated data: the construction of the "efficiency" color-magnitude diagram which will be used to calculate our experimental sensitivity for detecting microlensing in the LMC.

  11. Herschel celestial calibration sources: Four large main-belt asteroids as prime flux calibrators for the far-IR/sub-mm range

    CERN Document Server

    Mueller, T G; Nielbock, M; Lim, T; Teyssier, D; Olberg, M; Klaas, U; Linz, H; Altieri, B; Pearson, C; Bendo, G; Vilenius, E

    2013-01-01

    Celestial standards play a major role in observational astrophysics. They are needed to characterise the performance of instruments and are paramount for photometric calibration. During the Herschel Calibration Asteroid Preparatory Programme approximately 50 asteroids have been established as far-IR/sub-mm/mm calibrators for Herschel. The selected asteroids fill the flux gap between the sub-mm/mm calibrators Mars, Uranus and Neptune, and the mid-IR bright calibration stars. All three Herschel instruments observed asteroids for various calibration purposes, including pointing tests, absolute flux calibration, relative spectral response function, observing mode validation, and cross-calibration aspects. Here we present newly established models for the four large and well characterized main-belt asteroids (1) Ceres, (2) Pallas, (4) Vesta, and (21) Lutetia which can be considered as new prime flux calibrators. The relevant object-specific properties (size, shape, spin-properties, albedo, thermal properties) are w...

  12. Absolute spectral radiance responsivity calibration of sun photometers

    International Nuclear Information System (INIS)

    Sun photometers are designed to measure direct solar irradiance and diffused sky radiance for the purpose of atmospheric parameters characterization. A sun photometer is usually calibrated by using a lamp-illuminated integrating sphere source for its band-averaged radiance responsivity, which normally has an uncertainty of 3%-5% at present. Considering the calibration coefficients may also change with time, a regular high precision calibration is important to maintain data quality. In this paper, a tunable-laser-based facility for spectral radiance responsivity calibration has been developed at the Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences. A reference standard radiance radiometer, calibrated against cryogenic radiometer, is used to determine the radiance from a laser-illuminated integrating sphere source. Spectral radiance responsivity of CIMEL CE318-2 sun photometer is calibrated using this new calibration system with a combined standard uncertainty of about 0.8%. As a validation, the derived band-averaged radiance responsivity are compared to that from a Goddard Space Flight Center lamp-based sphere calibration and good agreements (difference <1.4%) are found from 675 to 1020 nm bands.

  13. TARGETLESS CAMERA CALIBRATION

    Directory of Open Access Journals (Sweden)

    L. Barazzetti

    2012-09-01

    Full Text Available In photogrammetry a camera is considered calibrated if its interior orientation parameters are known. These encompass the principal distance, the principal point position and some Additional Parameters used to model possible systematic errors. The current state of the art for automated camera calibration relies on the use of coded targets to accurately determine the image correspondences. This paper presents a new methodology for the efficient and rigorous photogrammetric calibration of digital cameras which does not require any longer the use of targets. A set of images depicting a scene with a good texture are sufficient for the extraction of natural corresponding image points. These are automatically matched with feature-based approaches and robust estimation techniques. The successive photogrammetric bundle adjustment retrieves the unknown camera parameters and their theoretical accuracies. Examples, considerations and comparisons with real data and different case studies are illustrated to show the potentialities of the proposed methodology.

  14. Landsat-8 Operational Land Imager Radiometric Calibration and Stability

    Directory of Open Access Journals (Sweden)

    Brian Markham

    2014-12-01

    Full Text Available The Landsat-8 Operational Land Imager (OLI was radiometrically calibrated prior to launch in terms of spectral radiance, using an integrating sphere source traceable to National Institute of Standards and Technology (NIST standards of spectral irradiance. It was calibrated on-orbit in terms of reflectance using diffusers characterized prior to launch using NIST traceable standards. The radiance calibration was performed with an uncertainty of ~3%; the reflectance calibration to an uncertainty of ~2%. On-orbit, multiple calibration techniques indicate that the sensor has been stable to better than 0.3% to date, with the exception of the shortest wavelength band, which has degraded about 1.0%. A transfer to orbit experiment conducted using the OLI’s heliostat-illuminated diffuser suggests that some bands increased in sensitivity on transition to orbit by as much as 5%, with an uncertainty of ~2.5%. On-orbit comparisons to other instruments and vicarious calibration techniques show the radiance (without a transfer to orbit adjustment, and reflectance calibrations generally agree with other instruments and ground measurements to within the uncertainties. Calibration coefficients are provided with the data products to convert to either radiance or reflectance units.

  15. 40 CFR 86.120-94 - Gas meter or flow instrumentation calibration; particulate, methanol and formaldehyde measurement.

    Science.gov (United States)

    2010-07-01

    ... laminar flow element or an NBS traceable flow calibration device is required as the standard device. (ii... agrees with the calibration measurement at the specified flow rates using the criteria of paragraph (a)(6... function of the calibration device flow measurement) from the calibration points to determine...

  16. The MINOS calibration detector

    International Nuclear Information System (INIS)

    This paper describes the MINOS calibration detector (CalDet) and the procedure used to calibrate it. The CalDet, a scaled-down but functionally equivalent model of the MINOS Far and Near detectors, was exposed to test beams in the CERN PS East Area during 2001-2003 to establish the response of the MINOS calorimeters to hadrons, electrons and muons in the range 0.2-10GeV/c. The CalDet measurements are used to fix the energy scale and constrain Monte Carlo simulations of MINOS

  17. Individual dosimetry and calibration

    International Nuclear Information System (INIS)

    In 1995 both the Individual Dosimetry and Calibration Sections worked under the condition of a status quo and concentrated fully on the routine part of their work. Nevertheless, the machine for printing the bar code which will be glued onto the film holder and hence identify the people when entering into high radiation areas was put into operation and most of the holders were equipped with the new identification. As far as the Calibration Section is concerned the project of the new source control system that is realized by the Technical Support Section was somewhat accelerated

  18. Calibration of PIXE yields using binary thin films on Si

    Energy Technology Data Exchange (ETDEWEB)

    Meersschaut, J., E-mail: Johan.Meersschaut@imec.be [Imec, Kapeldreef 75, B-3001 Leuven (Belgium); Carbonel, J.; Popovici, M. [Imec, Kapeldreef 75, B-3001 Leuven (Belgium); Zhao, Q.; Vantomme, A. [IKS, KU Leuven, Celestijnenlaan 200D, B-3001 Leuven (Belgium); Vandervorst, W. [Imec, Kapeldreef 75, B-3001 Leuven (Belgium); IKS, KU Leuven, Celestijnenlaan 200D, B-3001 Leuven (Belgium)

    2014-07-15

    We describe the use of binary thin films on Si to calibrate the yields in proton-induced X-ray emission (PIXE) measurements. Besides of the element to be calibrated, the standards also contain a common reference element. The incorporation of a common reference element allows one to eliminate errors in the accumulated beam charge during the calibration of the PIXE set-up. The binary calibration standards allow us to determine the response function with an accuracy close to 1%. As an example, we will perform the calibration for Fe and Co, and we will determine the Co concentration in Fe{sub 1−x}Co{sub x} thin films.

  19. A portable, automated, inexpensive mass and balance calibration system

    International Nuclear Information System (INIS)

    Reliable mass measurements are essential for a nuclear production facility or process control laboratory. DOE Order 5630.2 requires that traceable standards be used to calibrate and monitor equipment used for nuclear material measurements. To ensure the reliability of mass measurements and to comply with DOE traceability requirements, a portable, automated mass and balance calibration system is used at the Savannah River Plant. Automation is achieved using an EPSON HX-20 notebook computer, which can be operated via RS232C interfacing to electronic balances or function with manual data entry if computer interfacing is not feasible. This economical, comprehensive, user-friendly system has three main functions in a mass measurement control program (MMCP): balance certification, calibration of mass standards, and daily measurement of traceable standards. The balance certification program tests for accuracy, precision, sensitivity, linearity, and cornerloading versus specific requirements. The mass calibration program allows rapid calibration of inexpensive mass standards traceable to certified Class S standards. This MMCP permits daily measurement of traceable standards to monitor the reliability of balances during routine use. The automated system verifies balance calibration, stores results for future use, and provides a printed control chart of the stored data. Another feature of the system permits three different weighing routines that accommodate their need for varying degrees of reliability in routine weighing operations

  20. Near-Infrared Camera Calibration for Optical Surgical Navigation.

    Science.gov (United States)

    Cai, Ken; Yang, Rongqian; Lin, Qinyong; Liu, Sujuan; Chen, Huazhou; Ou, Shanxing; Huang, Wenhua; Zhou, Jing

    2016-03-01

    Near-infrared optical tracking devices, which are important components of surgical navigation systems, need to be calibrated for effective tracking. The calibration results has a direct influence on the tracking accuracy of an entire system. Therefore, the study of calibration techniques is of theoretical significance and practical value. In the present work, a systematic calibration method based on movable plates is established, which analyzes existing calibration theories and implements methods using calibration reference objects. First, the distortion model of near-infrared cameras (NICs) is analyzed in the implementation of this method. Second, the calibration images from different positions and orientations are used to establish the required linear equations. The initial values of the NIC parameters are calculated with the direct linear transformation method. Finally, the accurate internal and external parameters of the NICs are obtained by conducting nonlinear optimization. Analysis results show that the relative errors of the left and right NICs in the tracking system are 0.244 and 0.282 % for the focal lengths and 0.735 and 1.111 % for the principal points, respectively. The image residuals of the left and right image sets are both less than 0.01 pixel. The standard error of the calibration result is lower than 1, and the measurement error of the tracking system is less than 0.3 mm. The experimental data show that the proposed method of calibrating NICs is effective and can generate favorable calibration results. PMID:26728393

  1. A Calibrating Device for Rogowski Coil Development

    Institute of Scientific and Technical Information of China (English)

    LV Liang; LI Junhao; HUANG Jianjun; JI Shengchang; LI Yanming

    2007-01-01

    A calibrating device for the Rogowski coil is developed,which can be used to calibrate the Rogowski coil having a partial response time within tens of nanoseconds.Its key component is a step current generator,which can generate the output with a rise time of less than 2 ns and a duration of larger than 300 ns.The step current generator is composed by a pulse forming line(PFL)and a pulse transmission line(PTL).A TEM(transverse electromagnetic mode)coaxial measurement unit is used as PTL,and the coil to be calibrated and the referenced standard Rogowski coil can be fixed in the unit.The effect of the dimensions of the TEM unit is discussed theoretically as well as experimentally.

  2. Kinematic calibration of orthoglide-type mechanisms

    CERN Document Server

    Pashkevich, Anatoly; Wenger, Philippe

    2006-01-01

    The paper proposes a novel calibration approach for the Orthoglide-type mechanisms based on observations of the manipulator leg parallelism during mo-tions between the prespecified test postures. It employs a low-cost measuring system composed of standard comparator indicators attached to the universal magnetic stands. They are sequentially used for measuring the deviation of the relevant leg location while the manipulator moves the TCP along the Cartesian axes. Using the measured differences, the developed algorithm estimates the joint offsets that are treated as the most essential parameters to be adjusted. The sensitivity of the meas-urement methods and the calibration accuracy are also studied. Experimental re-sults are presented that demonstrate validity of the proposed calibration technique

  3. MODIS solar reflective calibration traceability

    Science.gov (United States)

    Xiong, Xiaoxiong; Butler, Jim

    2009-08-01

    Long-term climate data records often consist of observations made by multiple sensors. It is, therefore, extremely important to have instrument overlap, to be able to track instrument stability, to quantify measurement uncertainties, and to establish an absolute measurement scale traceable to the International System of Units (SI). The Moderate Resolution Imaging Spectroradiometer (MODIS) is a key instrument for both the Terra and Aqua missions, which were launched in December 1999 and May 2002, respectively. It has 20 reflective solar bands (RSB) with wavelengths from 0.41 to 2.2μm and observes the Earth at three nadir spatial resolutions: 0.25km, 0.5km, and 1km. MODIS RSB on-orbit calibration is reflectance based with reference to the bi-directional reflectance factor (BRF) of its on-board solar diffuser (SD). The SD BRF characterization was made pre-launch by the instrument vendor using reference samples traceable directly to the National Institute of Standards and Technology (NIST). On-orbit SD reflectance degradation is tracked by an on-board solar diffuser stability monitor (SDSM). This paper provides details of this calibration chain, from pre-launch to on-orbit operation, and associated uncertainty assessments. Using MODIS as an example, this paper also discusses challenges and key design requirements for future missions developed for accurate climate studies.

  4. MODIS Solar Reflective Calibration Traceability

    Science.gov (United States)

    Xiong, Xiaoxiong; Butler, Jim

    2009-01-01

    Long-term climate data records often consist of observations made by multiple sensors. It is, therefore, extremely important to have instrument overlap, to be able to track instrument stability, to quantify, measurement uncertainties, and to establish absolute scale traceable to the International System of Units (SI). The Moderate Resolution Imaging Spectroradiometer (MODIS) is a key instrument for both the Terra and Aqua missions, which were launched in December 1999 and May 2002, respectively. It has 20 reflective solar bands (RSB) with wavelengths from 0.41 to 2.2 micrometers and observes the Earth at three nadir spatial resolutions: 0.25km, 0.5km, and 1km. MODIS RSB on-orbit calibration is reflectance based with reference to the bidirectional reflectance factor (BRF) of its on-board solar diffuser (SD). The SD BRF characterization was made pre-launch by the instrument vendor using reference samples traceable directly to the National Institute of Standards and Technology (NIST). On-orbit SD reflectance degradation is tracked by an on-board solar diffuser monitor (SDSM). This paper provides details of this calibration chain, from prelaunch to on-orbit operation, and associated uncertainty assessments. Using MODIS as an example, this paper also discusses challenges and key design requirements for future missions developed for accurate climate studies.

  5. Thermal neutron calibration of a tritium extraction facility using the 6Li(n,t)4He/197Au(n,γ)198Au cross section ratio for standardization

    International Nuclear Information System (INIS)

    Absolute tritium activities in a neutron-activated metallic lithium samples have been measured by liquid scintillation methods to provide data needed for the determination of capture-to-fission ratios in fast breeder reactor spectra and for recent measurements of the 7Li(n,n't)4He cross section. The tritium extraction facility used for all these experiments has now been calibrated by measuring the 6Li(n,t)4He/197Au/n,γ)198Au activity ratio for thermal neutrons and comparing the result with the well-known cross sections. The calculated-to-measured activity ratio was found to be 1.033 +- 0.018. 2 figures, 20 tables

  6. Coordinate Standard Measurement Development

    Energy Technology Data Exchange (ETDEWEB)

    Hanshaw, R.A.

    2000-02-18

    A Shelton Precision Interferometer Base, which is used for calibration of coordinate standards, was improved through hardware replacement, software geometry error correction, and reduction of vibration effects. Substantial increases in resolution and reliability, as well as reduction in sampling time, were achieved through hardware replacement; vibration effects were reduced substantially through modification of the machine component dampening and software routines; and the majority of the machine's geometry error was corrected through software geometry error correction. Because of these modifications, the uncertainty of coordinate standards calibrated on this device has been reduced dramatically.

  7. Entropic calibration revisited

    Energy Technology Data Exchange (ETDEWEB)

    Brody, Dorje C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)]. E-mail: d.brody@imperial.ac.uk; Buckley, Ian R.C. [Centre for Quantitative Finance, Imperial College, London SW7 2AZ (United Kingdom); Constantinou, Irene C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom); Meister, Bernhard K. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)

    2005-04-11

    The entropic calibration of the risk-neutral density function is effective in recovering the strike dependence of options, but encounters difficulties in determining the relevant greeks. By use of put-call reversal we apply the entropic method to the time reversed economy, which allows us to obtain the spot price dependence of options and the relevant greeks.

  8. LOFAR Facet Calibration

    Science.gov (United States)

    van Weeren, R. J.; Williams, W. L.; Hardcastle, M. J.; Shimwell, T. W.; Rafferty, D. A.; Sabater, J.; Heald, G.; Sridhar, S. S.; Dijkema, T. J.; Brunetti, G.; Brüggen, M.; Andrade-Santos, F.; Ogrean, G. A.; Röttgering, H. J. A.; Dawson, W. A.; Forman, W. R.; de Gasperin, F.; Jones, C.; Miley, G. K.; Rudnick, L.; Sarazin, C. L.; Bonafede, A.; Best, P. N.; Bîrzan, L.; Cassano, R.; Chyży, K. T.; Croston, J. H.; Ensslin, T.; Ferrari, C.; Hoeft, M.; Horellou, C.; Jarvis, M. J.; Kraft, R. P.; Mevius, M.; Intema, H. T.; Murray, S. S.; Orrú, E.; Pizzo, R.; Simionescu, A.; Stroe, A.; van der Tol, S.; White, G. J.

    2016-03-01

    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction-dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction-dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at ∼ 5\\prime\\prime resolution, meeting the specifications of the LOFAR Tier-1 northern survey.

  9. LOFAR facet calibration

    CERN Document Server

    van Weeren, R J; Hardcastle, M J; Shimwell, T W; Rafferty, D A; Sabater, J; Heald, G; Sridhar, S S; Dijkema, T J; Brunetti, G; Brüggen, M; Andrade-Santos, F; Ogrean, G A; Röttgering, H J A; Dawson, W A; Forman, W R; de Gasperin, F; Jones, C; Miley, G K; Rudnick, L; Sarazin, C L; Bonafede, A; Best, P N; Bîrzan, L; Cassano, R; Chyży, K T; Croston, J H; Ensslin, T; Ferrari, C; Hoeft, M; Horellou, C; Jarvis, M J; Kraft, R P; Mevius, M; Intema, H T; Murray, S S; Orrú, E; Pizzo, R; Simionescu, A; Stroe, A; van der Tol, S; White, G J

    2016-01-01

    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at $\\sim$ 5arcsec resolu...

  10. Calibration Of Oxygen Monitors

    Science.gov (United States)

    Zalenski, M. A.; Rowe, E. L.; Mcphee, J. R.

    1988-01-01

    Readings corrected for temperature, pressure, and humidity of air. Program for handheld computer developed to ensure accuracy of oxygen monitors in National Transonic Facility, where liquid nitrogen stored. Calibration values, determined daily, based on entries of data on barometric pressure, temperature, and relative humidity. Output provided directly in millivolts.

  11. Commodity-Free Calibration

    Science.gov (United States)

    2008-01-01

    Commodity-free calibration is a reaction rate calibration technique that does not require the addition of any commodities. This technique is a specific form of the reaction rate technique, where all of the necessary reactants, other than the sample being analyzed, are either inherent in the analyzing system or specifically added or provided to the system for a reason other than calibration. After introduction, the component of interest is exposed to other reactants or flow paths already present in the system. The instrument detector records one of the following to determine the rate of reaction: the increase in the response of the reaction product, a decrease in the signal of the analyte response, or a decrease in the signal from the inherent reactant. With this data, the initial concentration of the analyte is calculated. This type of system can analyze and calibrate simultaneously, reduce the risk of false positives and exposure to toxic vapors, and improve accuracy. Moreover, having an excess of the reactant already present in the system eliminates the need to add commodities, which further reduces cost, logistic problems, and potential contamination. Also, the calculations involved can be simplified by comparison to those of the reaction rate technique. We conducted tests with hypergols as an initial investigation into the feasiblility of the technique.

  12. Measurement System & Calibration report

    DEFF Research Database (Denmark)

    Kock, Carsten Weber; Vesth, Allan

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  13. Calibration with Absolute Shrinkage

    DEFF Research Database (Denmark)

    Øjelund, Henrik; Madsen, Henrik; Thyregod, Poul

    2001-01-01

    In this paper, penalized regression using the L-1 norm on the estimated parameters is proposed for chemometric je calibration. The algorithm is of the lasso type, introduced by Tibshirani in 1996 as a linear regression method with bound on the absolute length of the parameters, but a modification...

  14. Pleiades Absolute Calibration : Inflight Calibration Sites and Methodology

    Science.gov (United States)

    Lachérade, S.; Fourest, S.; Gamet, P.; Lebègue, L.

    2012-07-01

    In-flight calibration of space sensors once in orbit is a decisive step to be able to fulfil the mission objectives. This article presents the methods of the in-flight absolute calibration processed during the commissioning phase. Four In-flight calibration methods are used: absolute calibration, cross-calibration with reference sensors such as PARASOL or MERIS, multi-temporal monitoring and inter-bands calibration. These algorithms are based on acquisitions over natural targets such as African deserts, Antarctic sites, La Crau (Automatic calibration station) and Oceans (Calibration over molecular scattering) or also new extra-terrestrial sites such as the Moon and selected stars. After an overview of the instrument and a description of the calibration sites, it is pointed out how each method is able to address one or several aspects of the calibration. We focus on how these methods complete each other in their operational use, and how they help building a coherent set of information that addresses all aspects of in-orbit calibration. Finally, we present the perspectives that the high level of agility of PLEIADES offers for the improvement of its calibration and a better characterization of the calibration sites.

  15. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten;

    2007-01-01

    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve the...... statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited for...

  16. Calibration of modified parallel-plate rheometer through calibrated oil and lattice Boltzmann simulation

    DEFF Research Database (Denmark)

    Ferraris, Chiara F; Geiker, Mette Rica; Martys, Nicos S;

    2007-01-01

    inapplicable here. This paper presents the analysis of a modified parallel plate rheometer for measuring cement mortar and propose a methodology for calibration using standard oils and numerical simulation of the flow. A lattice Boltzmann method was used to simulate the flow in the modified rheometer, thus...

  17. On the calibration of radiotherapy dosemeters in Australia

    International Nuclear Information System (INIS)

    Full text: Dosemeters for external beam radiotherapy are calibrated in Australia by ARPANSA, against the national primary standards of exposure and absorbed dose. The primary standards are free air chambers for exposure at low and medium energy X-rays, a graphite cavity chamber for exposure at 60Co, and a graphite calorimeter for absorbed dose at 60Co and high energy (MV) X -rays. Radiotherapy dosemeters are calibrated against these standards using a well documented formalism to provide calibration factors suitable for use with dosimetry protocols. A dosemeter usually comprises an ionization chamber connected to an independent electrometer. These are calibrated separately if possible. A combined calibration factor is reported together with the electrometer calibration factor (sensitivity). The dosimetry protocol used in radiotherapy centres in Australia and New Zealand is currently the simplified version of the IAEA TRS277 protocol, published by the New Zealand NRL and recommended by the ACPSEM. This protocol requires the use of an exposure or air kerma calibration factor at 60Co (Nx or Nk) to evaluate the absorbed dose to air calibration factor ND. The chamber is then placed in a water phantom with its centre displaced from the reference point by peff. ARPANSA can also supply calibration factors in absorbed dose to water (ND,w), as required as input to the new IAEA CoP. If an absorbed dose to water calibration factor is used by the radiotherapy centre, the chamber should be placed with its centre at the reference point in the water phantom. ARPANSA has for some years coordinated the participation of Australian radiotherapy centres in the IAEA TLD Quality Audit service. Note that this service does not represent a calibration and should not be referred to as such. The only calibration is that provided by ARPANSA for a reference dosemeter at each radiotherapy centre. As soon as the ANSTO SSDL is operational, calibrations of reference dosemeters will also be available

  18. In-situ Broadband Cryogenic Calibration for Two-port Superconducting Microwave Resonators

    CERN Document Server

    Yeh, Jen-Hao

    2012-01-01

    In this paper we introduce an improved microwave calibration method for use in a cryogenic environment, based on a traditional three-standard calibration, the Thru-Reflection-Line (TRL) calibration. The modified calibration method takes advantage of additional information from multiple measurements of an ensemble of realizations of a superconducting resonator, as a new pseudo-Open standard, to correct errors in the TRL calibration. We also demonstrate an experimental realization of this in-situ broadband cryogenic calibration system utilizing cryogenic switches. All calibration measurements are done in the same thermal cycle as the measurement of the resonator (requiring only an additional 20 minutes), thus avoiding 4 additional thermal cycles for traditional TRL calibration (which would require an additional 12 days). The experimental measurements on a wave chaotic microwave billiard verify that the new method significantly improves the measured scattering matrix of a high-quality-factor superconducting reso...

  19. In situ broadband cryogenic calibration for two-port superconducting microwave resonators.

    Science.gov (United States)

    Yeh, Jen-Hao; Anlage, Steven M

    2013-03-01

    We introduce an improved microwave calibration method for use in a cryogenic environment, based on a traditional three-standard calibration, the Thru-Reflect-Line (TRL) calibration. The modified calibration method takes advantage of additional information from multiple measurements of an ensemble of realizations of a superconducting resonator, as a new pseudo-Open standard, to correct errors in the TRL calibration. We also demonstrate an experimental realization of this in situ broadband cryogenic calibration system utilizing cryogenic switches. All calibration measurements are done in the same thermal cycle as the measurement of the resonator (requiring only an additional 20 min), thus avoiding 4 additional thermal cycles for traditional TRL calibration (which would require an additional 12 days). The experimental measurements on a wave-chaotic microwave billiard verify that the new method significantly improves the measured scattering matrix of a high-quality-factor superconducting resonator.

  20. Ophthalmic applicators: An overview of calibrations following the change to SI units

    International Nuclear Information System (INIS)

    Since the NIST dose to water standard for 90Sr/90Y ophthalmic applicators was introduced, numerous sources have undergone calibration either at NIST or at the University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL). From 1997 to 2008, 222 of these beta-emitting sources were calibrated at the UWADCL, and prior reference source strength values were available for 149 of these sources. A survey of UWADCL ophthalmic applicator calibrations is presented here, demonstrating an average discrepancy of -19% with a standard deviation of ±16% between prior reference values and the NIST-traceable UWADCL absorbed dose to water calibrations. Values ranged from -49% to +42%.

  1. Viscometry for liquids calibration of viscometers

    CERN Document Server

    Gupta, S V

    2014-01-01

    This book is written for scientists involved in the calibration of viscometers. A detailed description for stepping up procedures to establish the viscosity scale and obtaining sets of master viscometers is given in the book. Uncertainty considerations for standard oils of known viscosity are presented. The modern viscometers based on principles of  tuning fork, ultrasonic, PZT, plate waves, Love waves, micro-cantilever and vibration of optical fiber are discussed to inspire the reader to further research and to generate improved versions. The primary standard for viscosity is pure water. Measurements of its viscosity with accuracy/uncertainty achieved are described. The principles of rotational and oscillation viscometers are explained to enhance the knowledge in calibration work. Devices used for specific materials and viscosity in non SI units are discussed with respect to the need to correlate viscosity values obtained by various devices. The description of commercial viscometers meets the needs of the u...

  2. Polarimetric calibration of large mirrors

    CERN Document Server

    Ariste, A Lopez

    2015-01-01

    Aims: To propose a method for the polarimetric calibration of large astronomical mirrors that does not require use of special optical devices nor knowledge of the exact polarization properties of the calibration target. Methods: We study the symmetries of the Mueller matrix of mirrors to exploit them for polarimetric calibration under the assumptions that only the orientation of the linear polarization plane of the calibration target is known with certainty. Results: A method is proposed to calibrate the polarization effects of single astronomical mirrors by the observation of calibration targets with known orientation of the linear polarization. We study the uncertainties of the method and the signal-to-noise ratios required for an acceptable calibration. We list astronomical targets ready for the method. We finally extend the method to the calibration of two or more mirrors, in particular to the case when they share the same incidence plane.

  3. The Calibration Reference Data System

    Science.gov (United States)

    Greenfield, P.; Miller, T.

    2016-07-01

    We describe a software architecture and implementation for using rules to determine which calibration files are appropriate for calibrating a given observation. This new system, the Calibration Reference Data System (CRDS), replaces what had been previously used for the Hubble Space Telescope (HST) calibration pipelines, the Calibration Database System (CDBS). CRDS will be used for the James Webb Space Telescope (JWST) calibration pipelines, and is currently being used for HST calibration pipelines. CRDS can be easily generalized for use in similar applications that need a rules-based system for selecting the appropriate item for a given dataset; we give some examples of such generalizations that will likely be used for JWST. The core functionality of the Calibration Reference Data System is available under an Open Source license. CRDS is briefly contrasted with a sampling of other similar systems used at other observatories.

  4. Lidar calibration experiments

    DEFF Research Database (Denmark)

    Ejsing Jørgensen, Hans; Mikkelsen, T.; Streicher, J.;

    1997-01-01

    A series of atmospheric aerosol diffusion experiments combined with lidar detection was conducted to evaluate and calibrate an existing retrieval algorithm for aerosol backscatter lidar systems. The calibration experiments made use of two (almost) identical mini-lidar systems for aerosol cloud...... detection to test the reproducibility and uncertainty of lidars. Lidar data were obtained from both single-ended and double-ended Lidar configurations. A backstop was introduced in one of the experiments and a new method was developed where information obtained from the backstop can be used in the inversion...... algorithm. Independent in-situ aerosol plume concentrations were obtained from a simultaneous tracer gas experiment with SF6, and comparisons with the two lidars were made. The study shows that the reproducibility of the lidars is within 15%, including measurements from both sides of a plume...

  5. Optical tweezers absolute calibration

    CERN Document Server

    Dutra, R S; Neto, P A Maia; Nussenzveig, H M

    2014-01-01

    Optical tweezers are highly versatile laser traps for neutral microparticles, with fundamental applications in physics and in single molecule cell biology. Force measurements are performed by converting the stiffness response to displacement of trapped transparent microspheres, employed as force transducers. Usually, calibration is indirect, by comparison with fluid drag forces. This can lead to discrepancies by sizable factors. Progress achieved in a program aiming at absolute calibration, conducted over the past fifteen years, is briefly reviewed. Here we overcome its last major obstacle, a theoretical overestimation of the peak stiffness, within the most employed range for applications, and we perform experimental validation. The discrepancy is traced to the effect of primary aberrations of the optical system, which are now included in the theory. All required experimental parameters are readily accessible. Astigmatism, the dominant effect, is measured by analyzing reflected images of the focused laser spo...

  6. Astrid-2 SSC ASUMagnetic Calibration

    DEFF Research Database (Denmark)

    Primdahl, Fritz

    1997-01-01

    Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory.......Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory....

  7. Program Calibrates Strain Gauges

    Science.gov (United States)

    Okazaki, Gary D.

    1991-01-01

    Program dramatically reduces personnel and time requirements for acceptance tests of hardware. Data-acquisition system reads output from Wheatstone full-bridge strain-gauge circuit and calculates strain by use of shunt calibration technique. Program nearly instantaneously tabulates and plots strain data against load-cell outputs. Modified to acquire strain data for other specimens wherever full-bridge strain-gauge circuits used. Written in HP BASIC.

  8. Mesoscale hybrid calibration artifact

    Science.gov (United States)

    Tran, Hy D.; Claudet, Andre A.; Oliver, Andrew D.

    2010-09-07

    A mesoscale calibration artifact, also called a hybrid artifact, suitable for hybrid dimensional measurement and the method for make the artifact. The hybrid artifact has structural characteristics that make it suitable for dimensional measurement in both vision-based systems and touch-probe-based systems. The hybrid artifact employs the intersection of bulk-micromachined planes to fabricate edges that are sharp to the nanometer level and intersecting planes with crystal-lattice-defined angles.

  9. Electroweak Calibration of the Higgs Characterization Model

    CERN Document Server

    CERN. Geneva

    2015-01-01

    I will present the preliminary results of histogram fits using the Higgs Combine histogram fitting package. These fits can be used to estimate the effects of electroweak contributions to the p p -> H mu+ mu- Higgs production channel and calibrate Beyond Standard Model (BSM) simulations which ignore these effects. I will emphasize my findings' significance in the context of other research here at CERN and in the broader world of high energy physics.

  10. The Herschel PACS photometer calibration - A time dependent flux calibration for the PACS chopped point-source photometry AOT mode

    CERN Document Server

    Nielbock, Markus; Klaas, Ulrich; Altieri, Bruno; Balog, Zoltán; Billot, Nicolas; Linz, Hendrik; Okumura, Koryo; Sánchez-Portal, Miguel; Sauvage, Marc

    2013-01-01

    We present a flux calibration scheme for the PACS chopped point-source photometry observing mode based on the photometry of five stellar standard sources. This mode was used for science observations only early in the mission. Later, it was only used for pointing and flux calibration measurements. Its calibration turns this type of observation into fully validated data products in the Herschel Science Archive. Systematic differences in calibration with regard to the principal photometer observation mode, the scan map, are derived and amount to 5-6%. An empirical method to calibrate out an apparent response drift during the first 300 Operational Days is presented. The relative photometric calibration accuracy (repeatability) is as good as 1% in the blue and green band and up to 5% in the red band. Like for the scan map mode, inconsistencies among the stellar calibration models become visible and amount to 2% for the five standard stars used. The absolute calibration accuracy is therefore mainly limited by the m...

  11. Dosimetry and Calibration Section

    International Nuclear Information System (INIS)

    The two tasks of the Dosimetry and Calibration Section at CERN are the Individual Dosimetry Service which assures the personal monitoring of about 5000 persons potentially exposed to ionizing radiation at CERN, and the Calibration Laboratory which verifies all the instruments and monitors. This equipment is used by the sections of the RP Group for assuring radiation protection around CERN's accelerators, and by the Environmental Section of TISTE. In addition, nearly 250 electronic and 300 quartz fibre dosimeters, employed in operational dosimetry, are calibrated at least once a year. The Individual Dosimetry Service uses an extended database (INDOS) which contains information about all the individual doses ever received at CERN. For most of 1997 it was operated without the support of a database administrator as the technician who had assured this work retired. The Software Support Section of TIS-TE took over the technical responsibility of the database, but in view of the many other tasks of this Section and the lack of personnel, only a few interventions for solving immediate problems were possible

  12. Calibrating Historical IR Sensors Using GEO, and AVHRR Infrared Tropical Mean Calibration Models

    Science.gov (United States)

    Scarino, Benjamin; Doelling, David R.; Minnis, Patrick; Gopalan, Arun; Haney, Conor; Bhatt, Rajendra

    2014-01-01

    Long-term, remote-sensing-based climate data records (CDRs) are highly dependent on having consistent, wellcalibrated satellite instrument measurements of the Earth's radiant energy. Therefore, by making historical satellite calibrations consistent with those of today's imagers, the Earth-observing community can benefit from a CDR that spans a minimum of 30 years. Most operational meteorological satellites rely on an onboard blackbody and space looks to provide on-orbit IR calibration, but neither target is traceable to absolute standards. The IR channels can also be affected by ice on the detector window, angle dependency of the scan mirror emissivity, stray-light, and detector-to-detector striping. Being able to quantify and correct such degradations would mean IR data from any satellite imager could contribute to a CDR. Recent efforts have focused on utilizing well-calibrated modern hyper-spectral sensors to intercalibrate concurrent operational IR imagers to a single reference. In order to consistently calibrate both historical and current IR imagers to the same reference, however, another strategy is needed. Large, well-characterized tropical-domain Earth targets have the potential of providing an Earth-view reference accuracy of within 0.5 K. To that effort, NASA Langley is developing an IR tropical mean calibration model in order to calibrate historical Advanced Very High Resolution Radiometer (AVHRR) instruments. Using Meteosat-9 (Met-9) as a reference, empirical models are built based on spatially/temporally binned Met-9 and AVHRR tropical IR brightness temperatures. By demonstrating the stability of the Met-9 tropical models, NOAA-18 AVHRR can be calibrated to Met-9 by matching the AVHRR monthly histogram averages with the Met-9 model. This method is validated with ray-matched AVHRR and Met-9 biasdifference time series. Establishing the validity of this empirical model will allow for the calibration of historical AVHRR sensors to within 0.5 K, and thereby

  13. Calibration of Underwater Sound Transducers

    Directory of Open Access Journals (Sweden)

    H.R.S. Sastry

    1983-07-01

    Full Text Available The techniques of calibration of underwater sound transducers for farfield, near-field and closed environment conditions are reviewed in this paper .The design of acoustic calibration tank is mentioned. The facilities available at Naval Physical & Oceanographic Laboratory, Cochin for calibration of transducers are also listed.

  14. SWIR calibration of Spectralon reflectance factor

    Science.gov (United States)

    Georgiev, Georgi T.; Butler, James J.; Cooksey, Catherine; Ding, Leibo; Thome, Kurtis J.

    2011-11-01

    Satellite instruments operating in the reflective solar wavelength region require accurate and precise determination of the Bidirectional Reflectance Factor (BRF) of laboratory-based diffusers used in their pre-flight and on-orbit radiometric calibrations. BRF measurements are required throughout the reflected-solar spectrum from the ultraviolet through the shortwave infrared. Spectralon diffusers are commonly used as a reflectance standard for bidirectional and hemispherical geometries. The Diffuser Calibration Laboratory (DCaL) at NASA's Goddard Space Flight Center is a secondary calibration facility with reflectance measurements traceable to those made by the Spectral Tri-function Automated Reference Reflectometer (STARR) facility at the National Institute of Standards and Technology (NIST). For more than two decades, the DCaL has provided numerous NASA projects with BRF data in the ultraviolet (UV), visible (VIS) and the Near InfraRed (NIR) spectral regions. Presented in this paper are measurements of BRF from 1475 nm to 1625 nm obtained using an indium gallium arsenide detector and a tunable coherent light source. The sample was a 50.8 mm (2 in) diameter, 99% white Spectralon target. The BRF results are discussed and compared to empirically generated data from a model based on NIST certified values of 6°directional-hemispherical spectral reflectance factors from 900 nm to 2500 nm. Employing a new NIST capability for measuring bidirectional reflectance using a cooled, extended InGaAs detector, BRF calibration measurements of the same sample were also made using NIST's STARR from 1475 nm to 1625 nm at an incident angle of 0° and at viewing angle of 45°. The total combined uncertainty for BRF in this ShortWave Infrared (SWIR) range is less than 1%. This measurement capability will evolve into a BRF calibration service in SWIR region in support of NASA remote sensing missions.

  15. Calibration effects on orbit determination

    Science.gov (United States)

    Madrid, G. A.; Winn, F. B.; Zielenbach, J. W.; Yip, K. B.

    1974-01-01

    The effects of charged particle and tropospheric calibrations on the orbit determination (OD) process are analyzed. The calibration process consisted of correcting the Doppler observables for the media effects. Calibrated and uncalibrated Doppler data sets were used to obtain OD results for past missions as well as Mariner Mars 1971. Comparisons of these Doppler reductions show the significance of the calibrations. For the MM'71 mission, the media calibrations proved themselves effective in diminishing the overall B-plane error and reducing the Doppler residual signatures.

  16. Calibrating echelle spectrographs with Fabry-Perot etalons

    CERN Document Server

    Bauer, Florian F; Reiners, Ansgar

    2015-01-01

    Over the past decades hollow-cathode lamps have been calibration standards for spectroscopic measurements. Advancing to cm/s radial velocity precisions with the next generation of instruments requires more suitable calibration sources with more lines and less dynamic range problems. Fabry-Perot interferometers provide a regular and dense grid of lines and homogeneous amplitudes making them good candidates for next generation calibrators. We investigate the usefulness of Fabry-Perot etalons in wavelength calibration, present an algorithm to incorporate the etalon spectrum in the wavelength solution and examine potential problems. The quasi periodic pattern of Fabry-Perot lines is used along with a hollow-cathode lamp to anchor the numerous spectral features on an absolute scale. We test our method with the HARPS spectrograph and compare our wavelength solution to the one derived from a laser frequency comb. The combined hollow-cathode lamp/etalon calibration overcomes large distortion (50 m/s) in the wavelengt...

  17. A variable acceleration calibration system

    Science.gov (United States)

    Johnson, Thomas H.

    2011-12-01

    A variable acceleration calibration system that applies loads using gravitational and centripetal acceleration serves as an alternative, efficient and cost effective method for calibrating internal wind tunnel force balances. Two proof-of-concept variable acceleration calibration systems are designed, fabricated and tested. The NASA UT-36 force balance served as the test balance for the calibration experiments. The variable acceleration calibration systems are shown to be capable of performing three component calibration experiments with an approximate applied load error on the order of 1% of the full scale calibration loads. Sources of error are indentified using experimental design methods and a propagation of uncertainty analysis. Three types of uncertainty are indentified for the systems and are attributed to prediction error, calibration error and pure error. Angular velocity uncertainty is shown to be the largest indentified source of prediction error. The calibration uncertainties using a production variable acceleration based system are shown to be potentially equivalent to current methods. The production quality system can be realized using lighter materials and a more precise instrumentation. Further research is needed to account for balance deflection, forcing effects due to vibration, and large tare loads. A gyroscope measurement technique is shown to be capable of resolving the balance deflection angle calculation. Long term research objectives include a demonstration of a six degree of freedom calibration, and a large capacity balance calibration.

  18. Rectifying calibration error of Goldmann applanation tonometer is easy!

    Directory of Open Access Journals (Sweden)

    Nikhil S Choudhari

    2014-01-01

    Full Text Available Purpose: Goldmann applanation tonometer (GAT is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT. Materials and Methods: Twenty-nine slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland were included in this cross-sectional interventional pilot study. The technique of rectification of calibration error of the tonometer involved cleaning and lubrication of the instrument followed by alignment of weights when lubrication alone didn′t suffice. We followed the South East Asia Glaucoma Interest Group′s definition of calibration error tolerance (acceptable GAT calibration error within ±2, ±3 and ±4 mm Hg at the 0, 20 and 60-mm Hg testing levels, respectively. Results: Twelve out of 29 (41.3% GATs were out of calibration. The range of positive and negative calibration error at the clinically most important 20-mm Hg testing level was 0.5 to 20 mm Hg and -0.5 to -18 mm Hg, respectively. Cleaning and lubrication alone sufficed to rectify calibration error of 11 (91.6% faulty instruments. Only one (8.3% faulty GAT required alignment of the counter-weight. Conclusions: Rectification of calibration error of GAT is possible in-house. Cleaning and lubrication of GAT can be carried out even by eye care professionals and may suffice to rectify calibration error in the majority of faulty instruments. Such an exercise may drastically reduce the downtime of the Gold standard tonometer.

  19. NVLAP activities at Department of Defense calibration laboratories

    Energy Technology Data Exchange (ETDEWEB)

    Schaeffer, D.M. [Defense Nuclear Agency, Alexandria, VA (United States)

    1993-12-31

    There are 367 active radiological instrument calibration laboratories within the U.S. Department of Defense (DoD). Each of the four services in DoD manages, operates, and certifies the technical proficiency and competency of those laboratories under their cognizance. Each service has designated secondary calibration laboratories to trace all calibration source standards to the National Institute of Standards and Technology. Individual service radiological calibration programs and capabilities, present and future, are described, as well as the measurement quality assurance (MQA) processes for their traceability. National Voluntary Laboratory Accreditation Program (NVLAP) programs for dosimetry systems are briefly summarized. Planned NVLAP accreditation of secondary laboratories is discussed in the context of current technical challenges and future efforts.

  20. High-dose secondary calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Humphreys, J.C. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program.

  1. NIST Calibration of a Neutron Spectrometer ROSPEC.

    Science.gov (United States)

    Heimbach, Craig

    2006-01-01

    A neutron spectrometer was acquired for use in the measurement of National Institute of Standards and Technology neutron fields. The spectrometer included options for the measurement of low and high energy neutrons, for a total measurement range from 0.01 eV up to 17 MeV. The spectrometer was evaluated in calibration fields and was used to determine the neutron spectrum of an Americium-Beryllium neutron source. The calibration fields used included bare and moderated (252)Cf, monoenergetic neutron fields of 2.5 MeV and 14 MeV, and a thermal-neutron beam. Using the calibration values determined in this exercise, the spectrometer gives a good approximation of the neutron spectrum, and excellent values for neutron fluence, for all NIST calibration fields. The spectrometer also measured an Americium-Beryllium neutron field in a NIST exposure facility and determined the field quite well. The spectrometer measured scattering effects in neutron spectra which previously could be determined only by calculation or integral measurements.

  2. SAR antenna calibration techniques

    Science.gov (United States)

    Carver, K. R.; Newell, A. C.

    1978-01-01

    Calibration of SAR antennas requires a measurement of gain, elevation and azimuth pattern shape, boresight error, cross-polarization levels, and phase vs. angle and frequency. For spaceborne SAR antennas of SEASAT size operating at C-band or higher, some of these measurements can become extremely difficult using conventional far-field antenna test ranges. Near-field scanning techniques offer an alternative approach and for C-band or X-band SARs, give much improved accuracy and precision as compared to that obtainable with a far-field approach.

  3. Calibrated user-friendly reverse transcriptase-PCR assay

    DEFF Research Database (Denmark)

    Bor, M V; Sørensen, B S; Rammer, P;

    1998-01-01

    We report a competitive reverse transcriptase-PCR (RT-PCR) assay and a calibrated user-friendly RT-PCR assay (CURT-PCR) for epidermal growth factor receptor (EGFR) mRNA. A calibrator was prepared from isolated rat liver RNA, and the amount of EGFR mRNA was determined by competitive RT-PCR. In CURT-PCR......, a calibration curve was developed by plotting the ratio between the amount of PCR product originating from the calibrator and the RNA internal standard vs the amount of EGFR mRNA present in the calibrator. A fixed amount of RNA internal standard was coamplified with the EGFR mRNA present in the...... calibrator or in the sample, using the same primer set. The primers were chosen in regions of the EGFR mRNA that show 100% homology between human, rat, and mouse. The amount of EGFR in the unknown samples was calculated from the calibration curve based on the ratio between PCR product originating from the...

  4. Common Calibration Source for Monitoring Long-term Ozone Trends

    Science.gov (United States)

    Kowalewski, Matthew

    2004-01-01

    Accurate long-term satellite measurements are crucial for monitoring the recovery of the ozone layer. The slow pace of the recovery and limited lifetimes of satellite monitoring instruments demands that datasets from multiple observation systems be combined to provide the long-term accuracy needed. A fundamental component of accurately monitoring long-term trends is the calibration of these various instruments. NASA s Radiometric Calibration and Development Facility at the Goddard Space Flight Center has provided resources to minimize calibration biases between multiple instruments through the use of a common calibration source and standardized procedures traceable to national standards. The Facility s 50 cm barium sulfate integrating sphere has been used as a common calibration source for both US and international satellite instruments, including the Total Ozone Mapping Spectrometer (TOMS), Solar Backscatter Ultraviolet 2 (SBUV/2) instruments, Shuttle SBUV (SSBUV), Ozone Mapping Instrument (OMI), Global Ozone Monitoring Experiment (GOME) (ESA), Scanning Imaging SpectroMeter for Atmospheric ChartographY (SCIAMACHY) (ESA), and others. We will discuss the advantages of using a common calibration source and its effects on long-term ozone data sets. In addition, sphere calibration results from various instruments will be presented to demonstrate the accuracy of the long-term characterization of the source itself.

  5. Design, Construction and Calibration of a Solar Radiation Measuring Meter

    OpenAIRE

    Asiegbu, A. Daniel; Echeweozo, E. Odinakachi

    2014-01-01

    A digital solar radiation measuring instrument has been designed, constructed and calibrated. It incorporates a small rectangular silicon photocell as the sensor. On exposure to solar radiation, electromotive force which is proportional to radiation intensity is developed within the circuit. The device correlates voltage developed with available solar intensity. A standard solarimeter was therefore used to calibrate the device to translate the unit of its reading from Volt to Watt per square ...

  6. Calibration for Radiation Protection Equipment for the Measuring Airborne Tritium

    Institute of Scientific and Technical Information of China (English)

    CHEN; Xi-lin; SHEN; En-wei; WEI; Ke-xin; WANG; Kong-zhao; LI; Hou-wen; GE; Jian-an; LV; Xiao-xia

    2012-01-01

    <正>Monitoring airborne tritium is an important routine work in heavy water reactor nuclear power stations and the units related with tritium. Nowadays direct measuring instruments like hand carrying tritium monitors are more often used in routine workshop environment check. Need for calibrating such monitors was suggested. A trial work about the calibration for radiation protection equipment for measuring airborne tritium was carried out with a domestic standard EJ/T 1077-1998 equivalent that of IEC 710.

  7. Comparison of Blackbodies for Calibration of Infrared Ear Thermometers

    DEFF Research Database (Denmark)

    Pušnik, Igor; Clausen, Sønnik; Favreau, Jacques-Olivier;

    2011-01-01

    The article presents the results of the EURAMET Project No. 927 “Comparison of blackbodies for calibration of infrared ear thermometers (IRETs)”. The objective of the comparison was to determine the agreement of blackbodies used for the calibration of IRETs among European national laboratories....... To verify the accuracy of an IRET, a suitable blackbody (BB) is needed. Such a blackbody related to the EN standard, designed for the calibration of ear thermometers and immersed in a stirred water bath, was provided for the comparison by the pilot laboratory. The pilot provided also the transfer IRET...

  8. Use of Radiometrically Calibrated Flat-Plate Calibrators in Calibration of Radiation Thermometers

    Science.gov (United States)

    Cárdenas-García, D.; Méndez-Lango, E.

    2015-08-01

    Most commonly used, low-temperature, infrared thermometers have large fields of view sizes that make them difficult to be calibrated with narrow aperture blackbodies. Flat-plate calibrators with large emitting surfaces have been proposed for calibrating these infrared thermometers. Because the emissivity of the flat plate is not unity, its radiance temperature is wavelength dependent. For calibration, the wavelength pass band of the device under test should match that of the reference infrared thermometer. If the device under test and reference radiometer have different pass bands, then it is possible to calculate the corresponding correction if the emissivity of the flat plate is known. For example, a correction of at is required when calibrating a infrared thermometer with a "" radiometrically calibrated flat-plate calibrator. A method is described for using a radiometrically calibrated flat-plate calibrator that covers both cases of match and mismatch working wavelength ranges of a reference infrared thermometer and infrared thermometers to be calibrated with the flat-plate calibrator. Also, an application example is included in this paper.

  9. Approaches on calibration of bolometer and establishment of bolometer calibration device

    Science.gov (United States)

    Xia, Ming; Gao, Jianqiang; Ye, Jun'an; Xia, Junwen; Yin, Dejin; Li, Tiecheng; Zhang, Dong

    2015-10-01

    Bolometer is mainly used for measuring thermal radiation in the field of public places, labor hygiene, heating and ventilation and building energy conservation. The working principle of bolometer is under the exposure of thermal radiation, temperature of black absorbing layer of detector rise after absorption of thermal radiation, which makes the electromotive force produced by thermoelectric. The white light reflective layer of detector does not absorb thermal radiation, so the electromotive force produced by thermoelectric is almost zero. A comparison of electromotive force produced by thermoelectric of black absorbing layer and white reflective layer can eliminate the influence of electric potential produced by the basal background temperature change. After the electromotive force which produced by thermal radiation is processed by the signal processing unit, the indication displays through the indication display unit. The measurement unit of thermal radiation intensity is usually W/m2 or kW/m2. Its accurate and reliable value has important significance for high temperature operation, labor safety and hygiene grading management. Bolometer calibration device is mainly composed of absolute radiometer, the reference light source, electric measuring instrument. Absolute radiometer is a self-calibration type radiometer. Its working principle is using the electric power which can be accurately measured replaces radiation power to absolutely measure the radiation power. Absolute radiometer is the standard apparatus of laser low power standard device, the measurement traceability is guaranteed. Using the calibration method of comparison, the absolute radiometer and bolometer measure the reference light source in the same position alternately which can get correction factor of irradiance indication. This paper is mainly about the design and calibration method of the bolometer calibration device. The uncertainty of the calibration result is also evaluated.

  10. A Simple Accelerometer Calibrator

    Science.gov (United States)

    Salam, R. A.; Islamy, M. R. F.; Munir, M. M.; Latief, H.; Irsyam, M.; Khairurrijal

    2016-08-01

    High possibility of earthquake could lead to the high number of victims caused by it. It also can cause other hazards such as tsunami, landslide, etc. In that case it requires a system that can examine the earthquake occurrence. Some possible system to detect earthquake is by creating a vibration sensor system using accelerometer. However, the output of the system is usually put in the form of acceleration data. Therefore, a calibrator system for accelerometer to sense the vibration is needed. In this study, a simple accelerometer calibrator has been developed using 12 V DC motor, optocoupler, Liquid Crystal Display (LCD) and AVR 328 microcontroller as controller system. The system uses the Pulse Wave Modulation (PWM) form microcontroller to control the motor rotational speed as response to vibration frequency. The frequency of vibration was read by optocoupler and then those data was used as feedback to the system. The results show that the systems could control the rotational speed and the vibration frequencies in accordance with the defined PWM.

  11. Generic System for Remote Testing and Calibration of Measuring Instruments: Security Architecture

    Science.gov (United States)

    Jurčević, M.; Hegeduš, H.; Golub, M.

    2010-01-01

    Testing and calibration of laboratory instruments and reference standards is a routine activity and is a resource and time consuming process. Since many of the modern instruments include some communication interfaces, it is possible to create a remote calibration system. This approach addresses a wide range of possible applications and permits to drive a number of different devices. On the other hand, remote calibration process involves a number of security issues due to recommendations specified in standard ISO/IEC 17025, since it is not under total control of the calibration laboratory personnel who will sign the calibration certificate. This approach implies that the traceability and integrity of the calibration process directly depends on the collected measurement data. The reliable and secure remote control and monitoring of instruments is a crucial aspect of internet-enabled calibration procedure.

  12. Two-Step Camera Calibration Method Developed for Micro UAV'S

    Science.gov (United States)

    Gašparović, M.; Gajski, D.

    2016-06-01

    The development of unmanned aerial vehicles (UAVs) and continuous price reduction of unmanned systems attracted us to this research. Professional measuring systems are dozens of times more expensive and often heavier than "amateur", non-metric UAVs. For this reason, we tested the DJI Phantom 2 Vision Plus UAV. Phantom's smaller mass and velocity can develop less kinetic energy in relation to the professional measurement platforms, which makes it potentially less dangerous for use in populated areas. In this research, we wanted to investigate the ability of such non-metric UAV and find the procedures under which this kind of UAV may be used for the photogrammetric survey. It is important to emphasize that UAV is equipped with an ultra wide-angle camera with 14MP sensor. Calibration of such cameras is a complex process. In the research, a new two-step process is presented and developed, and the results are compared with standard one-step camera calibration procedure. Two-step process involves initially removed distortion on all images, and then uses these images in the phototriangulation with self-calibration. The paper presents statistical indicators which proved that the proposed two-step process is better and more accurate procedure for calibrating those types of cameras than standard one-step calibration. Also, we suggest two-step calibration process as the standard for ultra-wideangle cameras for unmanned aircraft.

  13. Photometric Calibration of the Supernova Legacy Survey Fields

    CERN Document Server

    Regnault, N; Guy, J; Sullivan, M; Cuillandre, J -C; Astier, P; Balland, C; Basa, S; Carlberg, R G; Fouchez, D; Hardin, D; Hook, I M; Howell, D A; Pain, R; Perrett, K; Pritchet, C J

    2009-01-01

    We present the photometric calibration of the Supernova Legacy Survey (SNLS) fields. The SNLS aims at measuring the distances to SNe Ia at (0.3calibration of the survey dominates the systematic uncertainty of the key measurement of the survey, namely the dark energy equation of state. The photometric calibration of the SNLS requires obtaining a uniform response across the imager, calibrating the science field stars in each survey band (SDSS-like ugriz bands) with respect to standards with known flux in the same bands, and binding the calibration to the UBVRI Landolt standards used to calibrate the nearby SNe from the literature necessary to produce cosmological constraints. The spatial non-uniformities of the imager photometric response are mapped using dithered observations of dense stellar fields. Photometric zero-points against Landolt standards are obtained. The linearity o...

  14. Absolute calibration of soft X-ray plates by using a self-calibrated X-ray bolometer

    International Nuclear Information System (INIS)

    The sensitivity of soft X-ray plates, Kodak 101-05 and ILFORD Q-PLATE at 1.1 keV was measured by using a self-calibrated X-ray bolometer as a standard detector for absolute measurements of soft X-rays. Cross-calibration was also made by using an X-ray sensitive diamond detector. Good agreement in the absolute sensitivity was obtained between the bolometer and the diamond detector. (author)

  15. On chromatic and geometrical calibration

    DEFF Research Database (Denmark)

    Folm-Hansen, Jørgen

    1999-01-01

    we present the implementation of a complete calibration method for an accurate colour texture measurement device called VMX2000, the calibration for uneven laser sheet illumination in a flow measuring system and the use of automatic detection of calibration targets for a DLT/warping in a 3D PIV......The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... of non-uniformity of the illumination of the image plane. Only the image deforming aberrations and the non-uniformity of illumination are included in the calibration models. The topics of the pinhole camera model and the extension to the Direct Linear Transform (DLT) are described. It is shown how...

  16. Calibration of high resolution digital camera based on different photogrammetric methods

    International Nuclear Information System (INIS)

    This paper presents method of calibrating high-resolution digital camera based on different configuration which comprised of stereo and convergent. Both methods are performed in the laboratory and in the field calibration. Laboratory calibration is based on a 3D test field where a calibration plate of dimension 0.4 m × 0.4 m with grid of targets at different height is used. For field calibration, it uses the same concept of 3D test field which comprised of 81 target points located on a flat ground and the dimension is 9 m × 9 m. In this study, a non-metric high resolution digital camera called Canon Power Shot SX230 HS was calibrated in the laboratory and in the field using different configuration for data acquisition. The aim of the calibration is to investigate the behavior of the internal digital camera whether all the digital camera parameters such as focal length, principal point and other parameters remain the same or vice-versa. In the laboratory, a scale bar is placed in the test field for scaling the image and approximate coordinates were used for calibration process. Similar method is utilized in the field calibration. For both test fields, the digital images were acquired within short period using stereo and convergent configuration. For field calibration, aerial digital images were acquired using unmanned aerial vehicle (UAV) system. All the images were processed using photogrammetric calibration software. Different calibration results were obtained for both laboratory and field calibrations. The accuracy of the results is evaluated based on standard deviation. In general, for photogrammetric applications and other applications the digital camera must be calibrated for obtaining accurate measurement or results. The best method of calibration depends on the type of applications. Finally, for most applications the digital camera is calibrated on site, hence, field calibration is the best method of calibration and could be employed for obtaining accurate

  17. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities; Metodos de calibracao e de intercomparacao de calibradores de dose utilizados em servicos de medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Costa, Alessandro Martins da

    1999-07-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  18. Calibration procedure for zenith plummets

    Directory of Open Access Journals (Sweden)

    Jelena GUČEVIĆ

    2013-09-01

    Full Text Available Zenith plummets are used mainly in applied geodesy, in civil engineering surveying, for materialization of the local vertical. The error of the vertical deflection of the instrument is directly transferred to the error of the observing construction. That is why a proper calibration procedure for the zenithlot is required. Metrological laboratory of the Faculty of Civil Engineering in Belgrade developed such calibration procedure. Here we present a mathematical model of the calibration and some selected results.

  19. Calibration procedure for zenith plummets

    OpenAIRE

    Jelena GUČEVIĆ; Delčev, Siniša; Vukan OGRIZOVIĆ

    2013-01-01

    Zenith plummets are used mainly in applied geodesy, in civil engineering surveying, for materialization of the local vertical. The error of the vertical deflection of the instrument is directly transferred to the error of the observing construction. That is why a proper calibration procedure for the zenithlot is required. Metrological laboratory of the Faculty of Civil Engineering in Belgrade developed such calibration procedure. Here we present a mathematical model of the calibration and som...

  20. Calibration of neutron albedo dosemeters.

    Science.gov (United States)

    Schwartz, R B; Eisenhauer, C M

    2002-01-01

    It is shown that by calibrating neutron albedo dosemeters under the proper conditions, two complicating effects will essentially cancel out, allowing accurate calibrations with no need for explicit corrections. The 'proper conditions' are: a large room (> or = 8 m on a side). use of a D2O moderated 252Cf source, and a source-to-phantom calibration distance of approximately 70 cm. PMID:12212898

  1. Summary of KOMPSAT-5 Calibration and Validation

    Science.gov (United States)

    Yang, D.; Jeong, H.; Lee, S.; Kim, B.

    2013-12-01

    Korean Multi-Purpose Satellite 5 (KOMPSAT-5), equipped with high resolution X-band (9.66 GHz) Synthetic Aperture Radar (SAR), is planning to be launched on August 22, 2013. With the satellite's primary mission objective being providing Geographical Information System (GIS), Ocean monitoring and Land management, and Disaster and ENvironment monitoring (GOLDEN), it is expected that its applications for scientific research on geographical processes will be extensive. In order to meet its mission objective, the KOMPSAT-5 will provide three different kinds of SAR imaging modes; High Resolution Mode (1 m resolution, 5 km swath), Standard Mode (3 m resolution, 30 km swath), and Wide Swath Mode (20 m resolution, 100 km swath). The KOMPSAT-5 will be operated in a 550 km sun-synchronous, dawn- dusk orbit with a 28-day ground repeat cycle providing valuable image information on Earth surface day-or-night and even in bad weather condition. After successful launch of the satellite, it will go through Launch and Early Operation (LEOP) and In-Orbit Testing (IOT) period about for 6 months to carry out various tests on satellite bus and payload systems. The satellite bus system will be tested during the first 3 weeks after the launch focusing on the Attitude and Orbit Control Subsystem (AOCS) and Integrated GPS Occultation Receiver (IGOR) calibration. With the completion of bus system test, the SAR payload system will be calibrated during initial In-Flight check period (11 weeks) by the joint effort of Thales Alenia Space Italy (TAS-I) and Korea Aerospace Research Institute (KARI). The pointing and relative calibration will be carried out during this period by analyzing the doppler frequency and antenna beam pattern of reflected microwave signal from selected regions with uniform backscattering coefficients (e.g. Amazon rainforest). A dedicated SAR calibration, called primary calibration, will be allocated at the end of LEOP for 12 weeks to perform thorough calibration activities

  2. Primary calibration of AM and PM noise measurements

    OpenAIRE

    Rubiola, Enrico

    2009-01-01

    This report describes a method for the primary calibration of phase noise and amplitude noise measurement systems. In the field of metrology, the term "primary" refers to a standard whose quantity value and measurement uncertainty are established without relation to another measurement standard for a quantity of the same kind; or to a procedure used to realize the definition of a measurement unit and obtain the quantity value and measurement uncertainty of a primary measurement standard. Acco...

  3. Intelligent electric energy meter GPS standard clock synchronization clock calibration technology development andapplication based on%基于GPS标准时钟的智能电能表时钟同步校准技术开发和应用

    Institute of Scientific and Technical Information of China (English)

    黄建硕; 李福东

    2015-01-01

    针对智能电能表时钟管理问题,分析了电能表时钟工作原理和管理标准,在现有用电信息采集系统(主站)中开发了时钟管理相关软件模块,搭建了时钟校对实验平台。通过实验测试和现场应用,实现了基于GPS标准时钟的智能电能表时钟同步校准,提高了电能计量的可靠性。%Aiming at the intelligent electrical energy table clock management problems,analyzes the working principle of the electric energy meter clock and management standards,in the existing electric information acquisition system(master station)in the development of the management software module clock,clock calibrationexperiment platform is built.Through the experimental test and field application,the intelligent electric energy meter GPS standard clock calibration based on clock synchronization,improve the reliability of electric energy metering.

  4. Calibration Techniques for VERITAS

    CERN Document Server

    Hanna, David

    2007-01-01

    VERITAS is an array of four identical telescopes designed for detecting and measuring astrophysical gamma rays with energies in excess of 100 GeV. Each telescope uses a 12 m diameter reflector to collect Cherenkov light from air showers initiated by incident gamma rays and direct it onto a `camera' comprising 499 photomultiplier tubes read out by flash ADCs. We describe here calibration methods used for determining the values of the parameters which are necessary for converting the digitized PMT pulses to gamma-ray energies and directions. Use of laser pulses to determine and monitor PMT gains is discussed, as are measurements of the absolute throughput of the telescopes using muon rings.

  5. RX130 Robot Calibration

    Science.gov (United States)

    Fugal, Mario

    2012-10-01

    In order to create precision magnets for an experiment at Oak Ridge National Laboratory, a new reverse engineering method has been proposed that uses the magnetic scalar potential to solve for the currents necessary to produce the desired field. To make the magnet it is proposed to use a copper coated G10 form, upon which a drill, mounted on a robotic arm, will carve wires. The accuracy required in the manufacturing of the wires exceeds nominal robot capabilities. However, due to the rigidity as well as the precision servo motor and harmonic gear drivers, there are robots capable of meeting this requirement with proper calibration. Improving the accuracy of an RX130 to be within 35 microns (the accuracy necessary of the wires) is the goal of this project. Using feedback from a displacement sensor, or camera and inverse kinematics it is possible to achieve this accuracy.

  6. Measuring Systems for Thermometer Calibration in Low-Temperature Range

    Science.gov (United States)

    Szmyrka-Grzebyk, A.; Lipiński, L.; Manuszkiewicz, H.; Kowal, A.; Grykałowska, A.; Jancewicz, D.

    2011-12-01

    The national temperature standard for the low-temperature range between 13.8033 K and 273.16 K has been established in Poland at the Institute of Low Temperature and Structure Research (INTiBS). The standard consists of sealed cells for realization of six fixed points of the International Temperature Scale of 1990 (ITS-90) in the low-temperature range, an adiabatic cryostat and Isotech water and mercury triple-point baths, capsule standard resistance thermometers (CSPRT), and AC and DC bridges with standard resistors for thermometers resistance measurements. INTiBS calibrates CSPRTs at the low-temperature fixed points with uncertainties less than 1 mK. In lower temperature range—between 2.5 K and about 25 K — rhodium-iron (RhFe) resistance thermometers are calibrated by comparison with a standard which participated in the EURAMET.T-K1.1 comparison. INTiBS offers a calibration service for industrial platinum resistance thermometers and for digital thermometers between 77 K and 273 K. These types of thermometers may be calibrated at INTiBS also in a higher temperature range up to 550°C. The Laboratory of Temperature Standard at INTiBS acquired an accreditation from the Polish Centre for Accreditation. A management system according to EN ISO/IEC 17025:2005 was established at the Laboratory and presented on EURAMET QSM Forum.

  7. Coincidence corrected efficiency calibration of Compton-suppressed HPGe detectors

    Energy Technology Data Exchange (ETDEWEB)

    Aucott, T.

    2015-04-20

    The authors present a reliable method to calibrate the full-energy efficiency and the coincidence correction factors using a commonly-available mixed source gamma standard. This is accomplished by measuring the peak areas from both summing and non-summing decay schemes and simultaneously fitting both the full-energy efficiency, as well as the total efficiency, as functions of energy. By using known decay schemes, these functions can then be used to provide correction factors for other nuclides not included in the calibration standard.

  8. Tectonic calibrations in molecular dating

    Institute of Scientific and Technical Information of China (English)

    Ullasa KODANDARAMAIAH

    2011-01-01

    Molecular dating techniques require the use of calibrations, which are usually fossil or geological vicariance-based.Fossil calibrations have been criticised because they result only in minimum age estimates. Based on a historical biogeographic perspective, Ⅰ suggest that vicariance-based calibrations are more dangerous. Almost all analytical methods in historical biogeography are strongly biased towards inferring vicariance, hence vicariance identified through such methods is unreliable. Other studies, especially of groups found on Gondwanan fragments, have simply assumed vicariance. Although it was previously believed that vicariance was the predominant mode of speciation, mounting evidence now indicates that speciation by dispersal is common, dominating vicariance in several groups. Moreover, the possibility of speciation having occurred before the said geological event cannot be precluded. Thus, geological calibrations can under- or overestimate times, whereas fossil calibrations always result in minimum estimates. Another major drawback of vicariant calibrations is the problem of circular reasoning when the resulting estimates are used to infer ages of biogeographic events. Ⅰ argue that fossil-based dating is a superior alternative to vicariance, primarily because the strongest assumption in the latter, that speciation was caused by the said geological process, is more often than not the most tenuous. When authors prefer to use a combination of fossil and vicariant calibrations, one suggestion is to report results both with and without inclusion of the geological constraints. Relying solely on vicariant calibrations should be strictly avoided.

  9. Secondary standard dosimetry laboratory at INFLPR

    Energy Technology Data Exchange (ETDEWEB)

    Scarlat, F.; Minea, R.; Scarisoreanu, A.; Badita, E.; Sima, E.; Dumitrascu, M.; Stancu, E.; Vancea, C., E-mail: scarlat.f@gmail.com [National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest (Romania)

    2011-07-01

    National Institute for Laser, Plasma and Radiation Physics (INFLPR) has constructed a High Energy Secondary Standard Dosimetry Laboratory SSDL-STARDOOR - for performing dosimetric calibrations according to ISO IEC SR/EN 17025:2005 standards. This is outfitted with UNIDOS Secondary Standard Dosimeter from PTW (Freiburg Physikalisch-Technische Werksttaten) calibrated at the PTB-Braunschweig (German Federal Institute of Physics and Metrology). A radiation beam of the quality of Q used by our laboratory as calibration source are provided by INFLPR 7 MeV electron beam linear accelerator mounted in our facility. (author)

  10. The Advanced LIGO Photon Calibrators

    CERN Document Server

    Karki, S; Kandhasamy, S; Abbott, B P; Abbott, T D; Anders, E H; Berliner, J; Betzwieser, J; Daveloza, H P; Cahillane, C; Canete, L; Conley, C; Gleason, J R; Goetz, E; Kissel, J S; Izumi, K; Mendell, G; Quetschke, V; Rodruck, M; Sachdev, S; Sadecki, T; Schwinberg, P B; Sottile, A; Wade, M; Weinstein, A J; West, M; Savage, R L

    2016-01-01

    The two interferometers of the Laser Interferometry Gravitaional-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events, and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as Photon Calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO Photon Calibrators that are currently providing fiducial displacements on the order of $10^{-18}$ m/$\\sqrt{\\textrm{Hz}}$ with accuracy and precision of better ...

  11. Antenna Calibration and Measurement Equipment

    Science.gov (United States)

    Rochblatt, David J.; Cortes, Manuel Vazquez

    2012-01-01

    A document describes the Antenna Calibration & Measurement Equipment (ACME) system that will provide the Deep Space Network (DSN) with instrumentation enabling a trained RF engineer at each complex to perform antenna calibration measurements and to generate antenna calibration data. This data includes continuous-scan auto-bore-based data acquisition with all-sky data gathering in support of 4th order pointing model generation requirements. Other data includes antenna subreflector focus, system noise temperature and tipping curves, antenna efficiency, reports system linearity, and instrument calibration. The ACME system design is based on the on-the-fly (OTF) mapping technique and architecture. ACME has contributed to the improved RF performance of the DSN by approximately a factor of two. It improved the pointing performances of the DSN antennas and productivity of its personnel and calibration engineers.

  12. Extending Sensor Calibration Intervals in Nuclear Power Plants

    Energy Technology Data Exchange (ETDEWEB)

    Coble, Jamie B.; Meyer, Ryan M.; Ramuhalli, Pradeep; Bond, Leonard J.; Shumaker, Brent; Hashemian, Hash

    2012-11-15

    Currently in the USA, sensor recalibration is required at every refueling outage, and it has emerged as a critical path item for shortening outage duration. International application of calibration monitoring, such as at the Sizewell B plant in UK, has shown that sensors may operate for eight years, or longer, within calibration tolerances. Online monitoring can be employed to identify those sensors which require calibration, allowing for calibration of only those sensors which need it. The US NRC accepted the general concept of online monitoring for sensor calibration monitoring in 2000, but no plants have been granted the necessary license amendment to apply it. This project addresses key issues in advanced recalibration methodologies and provides the science base to enable adoption of best practices for applying online monitoring, resulting in a public domain standardized methodology for sensor calibration interval extension. Research to develop this methodology will focus on three key areas: (1) quantification of uncertainty in modeling techniques used for calibration monitoring, with a particular focus on non-redundant sensor models; (2) accurate determination of acceptance criteria and quantification of the effect of acceptance criteria variability on system performance; and (3) the use of virtual sensor estimates to replace identified faulty sensors to extend operation to the next convenient maintenance opportunity.

  13. Comparison of calibrations of a well type ionization chamber between the IAEA and the SSDL of Finland

    International Nuclear Information System (INIS)

    Since 1996, the IAEA has maintained standards for Low Dose Rate (LDR) brachytherapy dosimetry. These standards consist of two 137Cs sources, calibrated at the National Institute of Standards and Technology (NIST), USA. As with all calibrations, maintaining our knowledge and confidence in of the standards at the highest possible level is essential. One way of verifying the quality of our calibrations is by means of a comparison with another SSDL. The purpose of this report is to describe such a comparison. A comparison was performed between the IAEA and the SSDL of Finland (STUK). A well type chamber, HDR 1000Plus (Standard Imaging, USA) was calibrated at the IAEA Dosimetry Laboratory and sent to STUK. After calibration by STUK the chamber was returned to the IAEA and a check of the chamber's response was made. This was done in order to verify that the chamber calibration had not been altered as a result of transportation

  14. Foreground-Induced Biases in CMB Polarimeter Self-Calibration

    CERN Document Server

    Abitbol, M H; Johnson, B R

    2015-01-01

    Precise polarisation measurements of the cosmic microwave background (CMB) require accurate knowledge of the instrument orientation relative to the sky frame used to define the cosmological Stokes parameters. Suitable celestial calibration sources that could be used to measure the polarimeter orientation angle are limited, so current experiments commonly `self-calibrate.' The self-calibration method exploits the theoretical fact that the $EB$ and $TB$ cross-spectra of the CMB vanish in the standard cosmological model, so any detected $EB$ and $TB$ signals must be due to systematic errors. However, this assumption neglects the fact that polarized Galactic foregrounds in a given portion of the sky may have non-zero $EB$ and $TB$ cross-spectra. If these foreground signals remain in the observations, then they will bias the self-calibrated telescope polarisation angle and produce a spurious $B$-mode signal. In this paper we estimate the foreground-induced bias for various instrument configurations and then expand...

  15. One step geometrical calibration method for optical coherence tomography

    International Nuclear Information System (INIS)

    We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery. (paper)

  16. Drift-insensitive distributed calibration of probe microscope scanner in nanometer range: Virtual mode

    Science.gov (United States)

    Lapshin, Rostislav V.

    2016-08-01

    A method of distributed calibration of a probe microscope scanner is suggested. The main idea consists in a search for a net of local calibration coefficients (LCCs) in the process of automatic measurement of a standard surface, whereby each point of the movement space of the scanner can be characterized by a unique set of scale factors. Feature-oriented scanning (FOS) methodology is used as a basis for implementation of the distributed calibration permitting to exclude in situ the negative influence of thermal drift, creep and hysteresis on the obtained results. Possessing the calibration database enables correcting in one procedure all the spatial systematic distortions caused by nonlinearity, nonorthogonality and spurious crosstalk couplings of the microscope scanner piezomanipulators. To provide high precision of spatial measurements in nanometer range, the calibration is carried out using natural standards - constants of crystal lattice. One of the useful modes of the developed calibration method is a virtual mode. In the virtual mode, instead of measurement of a real surface of the standard, the calibration program makes a surface image "measurement" of the standard, which was obtained earlier using conventional raster scanning. The application of the virtual mode permits simulation of the calibration process and detail analysis of raster distortions occurring in both conventional and counter surface scanning. Moreover, the mode allows to estimate the thermal drift and the creep velocities acting while surface scanning. Virtual calibration makes possible automatic characterization of a surface by the method of scanning probe microscopy (SPM).

  17. A calibrated Franklin chimes

    Science.gov (United States)

    Gonta, Igor; Williams, Earle

    1994-05-01

    Benjamin Franklin devised a simple yet intriguing device to measure electrification in the atmosphere during conditions of foul weather. He constructed a system of bells, one of which was attached to a conductor that was suspended vertically above his house. The device is illustrated in a well-known painting of Franklin (Cohen, 1985). The elevated conductor acquired a potential due to the electric field in the atmosphere and caused a brass ball to oscillate between two bells. The purpose of this study is to extend Franklin's idea by constructing a set of 'chimes' which will operate both in fair and in foul weather conditions. In addition, a mathematical relationship will be established between the frequency of oscillation of a metallic sphere in a simplified geometry and the potential on one plate due to the electrification of the atmosphere. Thus it will be possible to calibrate the 'Franklin Chimes' and to obtain a nearly instantaneous measurement of the potential of the elevated conductor in both fair and foul weather conditions.

  18. PV Cell and Module Calibrations at NREL

    Energy Technology Data Exchange (ETDEWEB)

    Emery, Keith

    2012-10-22

    NREL has equipment to measure any conceivable cell or module technology. The lack of standards for low concentration modules complicates matters. Spectrally adjustable simulators are critical for more than three junctions. NREL's 10-channel fiber optic simulator has shown that the light can be set for each junction within 1% of what it would be under the reference spectrum for up to a five-junction cell. Uncertainty in module simulators dominated by spatial nonuniformity for calibration labs. Manufacturers can mitigate this error by using matched reference modules instead of cells.

  19. Calibration of germanium detectors

    International Nuclear Information System (INIS)

    The process of determining the energy-dependent detection probability with measurements using Ge (Li) and high-grade germanium detectors is described. The paper explains which standards are best for a given purpose and given requirements as to accuracy, and how to assess measuring geometry variations and summation corrections. (DG)

  20. TOD to TTP calibration

    NARCIS (Netherlands)

    Bijl, P.; Reynolds, J.P.; Vos, W.K.; Hogervorst, M.A.; Fanning, J.D.

    2011-01-01

    The TTP (Targeting Task Performance) metric, developed at NVESD, is the current standard US Army model to predict EO/IR Target Acquisition performance. This model however does not have a corresponding lab or field test to empirically assess the performance of a camera system. The TOD (Triangle Orien

  1. 基于稳定竞争自适应重加权采样的光谱分析无标模型传递方法%Calibration Transfer without Standards for Spectral Analysis Based on Stability Competitive Adaptive Reweighted Sampling

    Institute of Scientific and Technical Information of China (English)

    张晓羽; 李庆波; 张广军

    2014-01-01

    A novel calibration transfer method based on stability competitive adaptive reweighted sampling (SCARS) was pro-posed in the present paper .An informative criterion ,i .e .the stability index ,defined as the absolute value of regression coeffi-cient divided by its standard deviation was used .And the root mean squared error of prediction (RMSEP) after transfer was also used .The wavelength variables which were important and insensitive to influence of measurement parameters were selected . And then the differences in responses of different instruments or measurement conditions for a specific sample were eliminated or reduced to improve the calibration transfer results .Moreover ,in the proposed method ,the spectral variables were compressed , making calibration transfer more stable .The application of the proposed method to calibration transfer of NIR analysis was eval-uated by analyzing the corn with different NIR spectrometers .The results showed that this method can well correct the differ-ence between instruments and improve the analytical accuracy .The transfer results obtained by the proposed method ,orthogonal signal correction (OSC) ,Monte Carlo uninformative variable elimination (MCUVE) and competitive adaptive reweighted sam-pling (CARS) ,respectively ,for corn with different NIR spectrometers indicated that the former gave the best analytical accura-cy ,and was effective for the spectroscopic data compression which can simplify and optimize the transfer process .%提出了一种基于稳定竞争自适应重加权采样(stability competitive adaptive reweighted sampling , SCARS)的无标模型传递方法。利用有用信息标准即稳定度指数(定义为回归系数除以其标准偏差的绝对值)和传递后的预测均方根误差(root mean squared error of prediction ,RMSEP),选择重要的、受测样参数影响不敏感的波长变量,能够消除或减少不同仪器或测量条件对样本信息反应差异,提高模

  2. Calibrating the PAU Survey's 46 Filters

    Science.gov (United States)

    Bauer, A.; Castander, F.; Gaztañaga, E.; Serrano, S.; Sevilla, N.; Tonello, N.; PAU Team

    2016-05-01

    The Physics of the Accelerating Universe (PAU) Survey, being carried out by several Spanish institutions, will image an area of 100-200 square degrees in 6 broad and 40 narrow band optical filters. The team is building a camera (PAUCam) with 18 CCDs, which will be installed in the 4 meter William Herschel Telescope at La Palma in 2013. The narrow band filters will each cover 100Å, with the set spanning 4500-8500Å. The broad band set will consist of standard ugriZy filters. The narrow band filters will provide low-resolution (R˜50) photometric "spectra" for all objects observed in the survey, which will reach a depth of ˜24 mag in the broad bands and ˜22.5 mag (AB) in the narrow bands. Such precision will allow for galaxy photometric redshift errors of 0.0035(1+z), which will facilitate the measurement of cosmological parameters with precision comparable to much larger spectroscopic and photometric surveys. Accurate photometric calibration of the PAU data is vital to the survey's science goals, and is not straightforward due to the large and unusual filter set. We outline the data management pipelines being developed for the survey, both for nightly data reduction and coaddition of multiple epochs, with emphasis on the photometric calibration strategies. We also describe the tools we are developing to test the quality of the reduction and calibration.

  3. Calibrating the Prominence Magnetometer (ProMag)

    Science.gov (United States)

    Fox, Lewis; Casini, R.

    2013-07-01

    The Prominence Magnetometer (ProMag) is a dual-channel, dual-beam, slit-scanning, full Stokes spectro-polarimeter designed by the High Altitude Observatory at the National Center for Atmospheric Research (HAO/NCAR) for the study of the magnetism of solar prominences and filaments. It was deployed in August 2009 at the 40 cm coronagraph of the Evans Solar Facility (ESF) of the National Solar Observatory on Sacramento Peak (NSO/SP). In its standard mode of operation it acquires spectro-polarimetric maps of solar targets simultaneously in the two chromospheric lines of He I at 587.6 nm and 1083.0 nm. Since August 2011 ProMag has operated in “patrol mode” with a dedicated observer. We aim to routinely measure the vector magnetic field in prominences. The electro-optic modulator and polarization analyzer are integrated into a single mechanical unit located at the coude feed of the telescope. This location was necessary for proper co-alignment of the dual beams, but complicates the precise polarimeter calibration necessary to achieve the sensitivity required for prominence measurements (calibration method for ProMag, using a polarizer and retarder at coronagraph prime focus. Calibrations are recorded before and after observations. We discuss the success of this method and its limitations.

  4. A critical comparison of systematic calibration protocols for activated sludge models: a SWOT analysis.

    Science.gov (United States)

    Sin, Gürkan; Van Hulle, Stijn W H; De Pauw, Dirk J W; van Griensven, Ann; Vanrolleghem, Peter A

    2005-07-01

    Modelling activated sludge systems has gained an increasing momentum after the introduction of activated sludge models (ASMs) in 1987. Application of dynamic models for full-scale systems requires essentially a calibration of the chosen ASM to the case under study. Numerous full-scale model applications have been performed so far which were mostly based on ad hoc approaches and expert knowledge. Further, each modelling study has followed a different calibration approach: e.g. different influent wastewater characterization methods, different kinetic parameter estimation methods, different selection of parameters to be calibrated, different priorities within the calibration steps, etc. In short, there was no standard approach in performing the calibration study, which makes it difficult, if not impossible, to (1) compare different calibrations of ASMs with each other and (2) perform internal quality checks for each calibration study. To address these concerns, systematic calibration protocols have recently been proposed to bring guidance to the modeling of activated sludge systems and in particular to the calibration of full-scale models. In this contribution four existing calibration approaches (BIOMATH, HSG, STOWA and WERF) will be critically discussed using a SWOT (Strengths, Weaknesses, Opportunities, Threats) analysis. It will also be assessed in what way these approaches can be further developed in view of further improving the quality of ASM calibration. In this respect, the potential of automating some steps of the calibration procedure by use of mathematical algorithms is highlighted.

  5. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-07-01

    Accurate solar radiation data sets are critical to reducing the expenses associated with mitigating performance risk for solar energy conversion systems, and they help utility planners and grid system operators understand the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of calibration methodologies and the resulting calibration responsivities provided by radiometric calibration service providers such as the National Renewable Energy Laboratory (NREL) and manufacturers of radiometers. Some of these radiometers are calibrated indoors, and some are calibrated outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The reference radiometer calibrations are traceable to the World Radiometric Reference. These different methods of calibration demonstrated 1% to 2% differences in solar irradiance measurement. Analyzing these values will ultimately assist in determining the uncertainties of the radiometer data and will assist in developing consensus on a standard for calibration.

  6. Calibration of High Frequency MEMS Microphones

    Science.gov (United States)

    Shams, Qamar A.; Humphreys, William M.; Bartram, Scott M.; Zuckewar, Allan J.

    2007-01-01

    calibrate these microphones at frequencies up to 80 kHz. The technique relied on the use of a random, ultrasonic broadband centrifugal sound source located in a small anechoic chamber. Phase calibrations of the MEMS microphones were derived from cross spectral phase comparisons between the reference and test substitution microphones and an adjacent and invariant grazing-incidence 1/8-inch standard microphone.

  7. Objective calibration of regional climate models

    Science.gov (United States)

    Bellprat, O.; Kotlarski, S.; Lüthi, D.; SchäR, C.

    2012-12-01

    methodology is effective and objective. It is argued that objective calibration is an attractive tool and could become standard procedure after introducing new model implementations, or after a spatial transfer of a regional climate model. Objective calibration of parameterizations with regional models could also serve as a strategy toward improving parameterization packages of global climate models.

  8. Mexican national pyronometer network calibration

    Science.gov (United States)

    VAldes, M.; Villarreal, L.; Estevez, H.; Riveros, D.

    2013-12-01

    In order to take advantage of the solar radiation as an alternate energy source it is necessary to evaluate the spatial and temporal availability. The Mexican National Meterological Service (SMN) has a network with 136 meteorological stations, each coupled with a pyronometer for measuring the global solar radiation. Some of these stations had not been calibrated in several years. The Mexican Department of Energy (SENER) in order to count on a reliable evaluation of the solar resource funded this project to calibrate the SMN pyrometer network and validate the data. The calibration of the 136 pyronometers by the intercomparison method recommended by the World Meterological Organization (WMO) requires lengthy observations and specific environmental conditions such as clear skies and a stable atmosphere, circumstances that determine the site and season of the calibration. The Solar Radiation Section of the Instituto de Geofísica of the Universidad Nacional Autónoma de México is a Regional Center of the WMO and is certified to carry out the calibration procedures and emit certificates. We are responsible for the recalibration of the pyronometer network of the SMN. A continuous emission solar simulator with exposed areas with 30cm diameters was acquired to reduce the calibration time and not depend on atmospheric conditions. We present the results of the calibration of 10 thermopile pyronometers and one photovoltaic cell by the intercomparison method with more than 10000 observations each and those obtained with the solar simulator.

  9. Regression calibration with more surrogates than mismeasured variables

    KAUST Repository

    Kipnis, Victor

    2012-06-29

    In a recent paper (Weller EA, Milton DK, Eisen EA, Spiegelman D. Regression calibration for logistic regression with multiple surrogates for one exposure. Journal of Statistical Planning and Inference 2007; 137: 449-461), the authors discussed fitting logistic regression models when a scalar main explanatory variable is measured with error by several surrogates, that is, a situation with more surrogates than variables measured with error. They compared two methods of adjusting for measurement error using a regression calibration approximate model as if it were exact. One is the standard regression calibration approach consisting of substituting an estimated conditional expectation of the true covariate given observed data in the logistic regression. The other is a novel two-stage approach when the logistic regression is fitted to multiple surrogates, and then a linear combination of estimated slopes is formed as the estimate of interest. Applying estimated asymptotic variances for both methods in a single data set with some sensitivity analysis, the authors asserted superiority of their two-stage approach. We investigate this claim in some detail. A troubling aspect of the proposed two-stage method is that, unlike standard regression calibration and a natural form of maximum likelihood, the resulting estimates are not invariant to reparameterization of nuisance parameters in the model. We show, however, that, under the regression calibration approximation, the two-stage method is asymptotically equivalent to a maximum likelihood formulation, and is therefore in theory superior to standard regression calibration. However, our extensive finite-sample simulations in the practically important parameter space where the regression calibration model provides a good approximation failed to uncover such superiority of the two-stage method. We also discuss extensions to different data structures.

  10. Calibration of radiation measuring equipment for use in nuclear power plants and laboratories for quality assurance by KWU

    International Nuclear Information System (INIS)

    The calibration described is carried out for equipment developed by KWU for use in controlled areas and for radioactivity monitoring in nuclear power plants. For the design approval, these initial calibrations are made. Except for iodine measuring kits, the water, noble gas and aerosol monitoring equipment is calibrated integrally. PTB standards are used for this purpose, and special calibrations are carried out by KWU, especially of laboratory equipment including γ-γ concidences, such as borehole detectors and anticompton measuring kits. (DG)

  11. The Landsat Data Continuity Mission Operational Land Imager (OLI) Radiometric Calibration

    Science.gov (United States)

    Markham, Brian L.; Dabney, Philip W.; Murphy-Morris, Jeanine E.; Knight, Edward J.; Kvaran, Geir; Barsi, Julia A.

    2010-01-01

    The Operational Land Imager (OLI) on the Landsat Data Continuity Mission (LDCM) has a comprehensive radiometric characterization and calibration program beginning with the instrument design, and extending through integration and test, on-orbit operations and science data processing. Key instrument design features for radiometric calibration include dual solar diffusers and multi-lamped on-board calibrators. The radiometric calibration transfer procedure from NIST standards has multiple checks on the radiometric scale throughout the process and uses a heliostat as part of the transfer to orbit of the radiometric calibration. On-orbit lunar imaging will be used to track the instruments stability and side slither maneuvers will be used in addition to the solar diffuser to flat field across the thousands of detectors per band. A Calibration Validation Team is continuously involved in the process from design to operations. This team uses an Image Assessment System (IAS), part of the ground system to characterize and calibrate the on-orbit data.

  12. Surface dose rate calibration of Sr-90 plane ophthalmic applicators

    International Nuclear Information System (INIS)

    Calibration of an imported strontium-90 ophthalmic applicator at the U.S. National Bureau of Standards (now the National Institute of Standards and Technology) has disclose a significant discrepancy in dose rate calibration (32%-35%) with that quoted by the manufacturer. The University of Wisconsin has investigated this discrepancy and found that both laboratories use similar techniques and a version of the Bragg-Gray equation to yield dose rate estimates. Experimental results indicate a strong relationship between the size of the collecting electrode used in the extrapolation chamber and the resulting estimate of absorbed dose rate. Calibration of these applicators is reviewed and suggestions for improvement and further research are proposed

  13. A New Measurement Principle for Determining the Polarization Direction of Calibration Transponder Antennas

    OpenAIRE

    Björn J. Döring; Schwerdt, Marco

    2016-01-01

    Polarimetric synthetic aperture radar (SAR) systems require verification and calibration of their polarimetric performance after launch. Dual-antenna calibration transponders with freely rotatable, linearly-polarized antennas are one of the most common calibration targets for this task. Because the transponders are used as a measurement standard, any misorientation of the polarization direction of the transponder antennas leads to characterization errors for the SAR instrument. In this paper,...

  14. Photometric Calibration of the [$\\alpha$/Fe] Element: I. Calibration with UBV Photometry

    CERN Document Server

    Karaali, S; Bilir, S

    2016-01-01

    We present the calibration of the [$\\alpha$/Fe] element in terms of the ultra-violet excess for 469 dwarf stars with $0.325<(B-V)_0 \\leq 0.775$ mag corresponding the spectral type range F0-K2. The star sample is separated into nine sub-samples with equal range in $(B-V)_0$ colour, $\\Delta(B-V)_0=0.05$ mag, and a third degree polynomial is fitted to each dataset. Our calibrations provide [$\\alpha$/Fe] elements in the range [0.0, 0.4]. We applied the procedure to two sets of field stars and two sets of clusters. The mean and the corresponding standard deviation of the residuals for 43 field stars taken from the Hypatia catalogue are [$\\alpha$/Fe]=-0.090 and $\\sigma =0.102$ dex, while for the 39 ones taken from the same catalogue of stars used in the calibration are [$\\alpha$/Fe]=-0.009 and $\\sigma = 0.079$ dex, respectively. We showed that the differences between the mean of the residuals and standard deviations for two sets of clusters ([$\\alpha$/Fe]=0.073 and $\\sigma = 0.91$ dex; [$\\alpha$/Fe]=-0.012 and $...

  15. Calibration of "Babyline" RP instruments

    CERN Multimedia

    2015-01-01

      If you have old RP instrumentation of the “Babyline” type, as shown in the photo, please contact the Radiation Protection Group (Joffrey Germa, 73171) to have the instrument checked and calibrated. Thank you. Radiation Protection Group

  16. Calibration in quantitative analysis Part 1. General considerations

    NARCIS (Netherlands)

    Agterdenbos, J.

    1979-01-01

    Mathematical procedures for calibration require assumptions to be made, e.g. the homogeneity of variances and the mathematical relationship between the analyte content x and the signal y. Little is known about the magnitude of errors arising from incorrect assumptions. The variation of the standard

  17. Electro-optical equivalent calibration technology for high-energy laser energy meters

    Science.gov (United States)

    Wei, Ji Feng; Chang, Yan; Sun, Li Qun; Zhang, Kai; Hu, Xiao Yang; Zhang, Wei

    2016-04-01

    Electro-optical equivalent calibration with high calibration power and high equivalence is particularly well-suited to the calibration of high-energy laser energy meters. A large amount of energy is reserved during this process, however, which continues to radiate after power-off. This study measured the radiation efficiency of a halogen tungsten lamp during power-on and after power-off in order to calculate the total energy irradiated by a lamp until the high-energy laser energy meter reaches thermal equilibrium. A calibration system was designed based on the measurement results, and the calibration equivalence of the system was analyzed in detail. Results show that measurement precision is significantly affected by the absorption factor of the absorption chamber and by heat loss in the energy meter. Calibration precision is successfully improved by enhancing the equivalent power and reducing power-on time. The electro-optical equivalent calibration system, measurement uncertainty of which was evaluated as 2.4% (k = 2), was used to calibrate a graphite-cone-absorption-cavity absolute energy meter, yielding a calibration coefficient of 1.009 and measurement uncertainty of 3.5% (k = 2). A water-absorption-type high-energy laser energy meter with measurement uncertainty of 4.8% (k = 2) was considered the reference standard, and compared to the energy meter calibrated in this study, yielded a correction factor of 0.995 (standard deviation of 1.4%).

  18. Electro-optical equivalent calibration technology for high-energy laser energy meters.

    Science.gov (United States)

    Wei, Ji Feng; Chang, Yan; Sun, Li Qun; Zhang, Kai; Hu, Xiao Yang; Zhang, Wei

    2016-04-01

    Electro-optical equivalent calibration with high calibration power and high equivalence is particularly well-suited to the calibration of high-energy laser energy meters. A large amount of energy is reserved during this process, however, which continues to radiate after power-off. This study measured the radiation efficiency of a halogen tungsten lamp during power-on and after power-off in order to calculate the total energy irradiated by a lamp until the high-energy laser energy meter reaches thermal equilibrium. A calibration system was designed based on the measurement results, and the calibration equivalence of the system was analyzed in detail. Results show that measurement precision is significantly affected by the absorption factor of the absorption chamber and by heat loss in the energy meter. Calibration precision is successfully improved by enhancing the equivalent power and reducing power-on time. The electro-optical equivalent calibration system, measurement uncertainty of which was evaluated as 2.4% (k = 2), was used to calibrate a graphite-cone-absorption-cavity absolute energy meter, yielding a calibration coefficient of 1.009 and measurement uncertainty of 3.5% (k = 2). A water-absorption-type high-energy laser energy meter with measurement uncertainty of 4.8% (k = 2) was considered the reference standard, and compared to the energy meter calibrated in this study, yielded a correction factor of 0.995 (standard deviation of 1.4%).

  19. Pressures Detector Calibration and Measurement

    CERN Document Server

    AUTHOR|(CDS)2156315

    2016-01-01

    This is report of my first and second projects (of 3) in NA61. I did data taking and analysis in order to do calibration of pressure detectors and verified it. I analyzed the data by ROOT software using the C ++ programming language. The first part of my project was determination of calibration factor of pressure sensors. Based on that result, I examined the relation between pressure drop, gas flow rate of in paper filter and its diameter.

  20. Beam Imaging and Luminosity Calibration

    CERN Document Server

    Klute, Markus; Salfeld-Nebgen, Jakob

    2016-01-01

    We discuss a method to reconstruct two-dimensional proton bunch densities using vertex distributions accumulated during LHC beam-beam scans. The $x$-$y$ correlations in the beam shapes are studied and an alternative luminosity calibration technique is introduced. We demonstrate the method on simulated beam-beam scans and estimate the uncertainty on the luminosity calibration associated to the beam-shape reconstruction to be below 1\\%.

  1. WFC3: UVIS Dark Calibration

    Science.gov (United States)

    Bourque, Matthew; Biretta, John A.; Anderson, Jay; Baggett, Sylvia M.; Gunning, Heather C.; MacKenty, John W.

    2014-06-01

    Wide Field Camera 3 (WFC3), a fourth-generation imaging instrument on board the Hubble Space Telescope (HST), has exhibited excellent performance since its installation during Servicing Mission 4 in May 2009. The UVIS detector, comprised of two e2v CCDs, is one of two channels available on WFC3 and is named for its ultraviolet and visible light sensitivity. We present the various procedures and results of the WFC3/UVIS dark calibration, which monitors the health and stability of the UVIS detector, provides characterization of hot pixels and dark current, and produces calibration files to be used as a correction for dark current in science images. We describe the long-term growth of hot pixels and the impacts that UVIS Charge Transfer Efficiency (CTE) losses, postflashing, and proximity to the readout amplifiers have on the population. We also discuss the evolution of the median dark current, which has been slowly increasing since the start of the mission and is currently ~6 e-/hr/pix, averaged across each chip. We outline the current algorithm for creating UVIS dark calibration files, which includes aggressive cosmic ray masking, image combination, and hot pixel flagging. Calibration products are available to the user community, typically 3-5 days after initial processing, through the Calibration Database System (CDBS). Finally, we discuss various improvements to the calibration and monitoring procedures. UVIS dark monitoring will continue throughout and beyond HST’s current proposal cycle.

  2. Calibration and evaluation of the quality assurance during 1996 at the National Laboratory for ionising radiation

    International Nuclear Information System (INIS)

    The Swedish Radiation Protection Institute is the National Laboratory for the dosimetric quantities kerma, absorbed dose and dose equivalent. The activity is based on established calibration procedures and a QA program for the used standards. This report gives a brief summary of the calibrations performed during 1996 and a more detailed description and analysis of the QA this year. 10 figs, 9 tabs

  3. Two Methods for Self Calibration of Digital Camera

    Science.gov (United States)

    Sampath, A.; Moe, D.; Christopherson, J.

    2012-07-01

    Photogrammetric mapping using Commercial of the Shelf (COTS) cameras is becoming more popular. Their popularity is augmented by the increasing use of Unmanned Aerial Vehicles (UAV) as a platform for mapping. The mapping precision of these methods can be increased by using a calibrated camera. The USGS/EROS has developed an inexpensive, easy to use method, particularly for calibrating short focal length cameras. The method builds on a self-calibration procedure developed for the USGS EROS Data Center by Pictometry (and augmented by Dr. C.S Fraser), that uses a series of coded targets. These coded targets form different patterns that are imaged from nine different locations with differing camera orientations. A free network solution using collinearity equations is used to determine the calibration parameters. For the smaller focal length COTS cameras, the USGS has developed a procedure that uses a small prototype box that contains these coded targets. The design of the box is discussed, along with best practices for calibration procedure. Results of calibration parameters obtained using the box are compared with the parameters obtained using more established standard procedures.

  4. Comparative Analysis of Different LIDAR System Calibration Techniques

    Science.gov (United States)

    Miller, M.; Habib, A.

    2016-06-01

    With light detection and ranging (LiDAR) now being a crucial tool for engineering products and on the fly spatial analysis, it is necessary for the user community to have standardized calibration methods. The three methods in this study were developed and proven by the Digital Photogrammetry Research Group (DPRG) for airborne LiDAR systems and are as follows; Simplified, Quasi-Rigorous, and Rigorous. In lieu of using expensive control surfaces for calibration, these methods compare overlapping LiDAR strips to estimate the systematic errors. These systematic errors are quantified by these methods and include the lever arm biases, boresight biases, range bias and scan angle scale bias. These three methods comprehensively represent all of the possible flight configurations and data availability and this paper will test the limits of the method with the most assumptions, the simplified calibration, by using data that violates the assumptions it's math model is based on and compares the results to the quasi-rigorous and rigorous techniques. The overarching goal is to provide a LiDAR system calibration that does not require raw measurements which can be carried out with minimal control and flight lines to reduce costs. This testing is unique because the terrain used for calibration does not contain gable roofs, all other LiDAR system calibration testing and development has been done with terrain containing features with high geometric integrity such as gable roofs.

  5. Calibrating echelle spectrographs with Fabry-Pérot etalons

    Science.gov (United States)

    Bauer, F. F.; Zechmeister, M.; Reiners, A.

    2015-09-01

    Context. Over the past decades hollow-cathode lamps have been calibration standards for spectroscopic measurements. Advancing to cm/s radial velocity precisions with the next generation of instruments requires more suitable calibration sources with more lines and fewer dynamic range problems. Fabry-Pérot interferometers provide a regular and dense grid of lines and homogeneous amplitudes, which makes them good candidates for next-generation calibrators. Aims: We investigate the usefulness of Fabry-Pérot etalons in wavelength calibration, present an algorithm to incorporate the etalon spectrum in the wavelength solution, and examine potential problems. Methods: The quasi-periodic pattern of Fabry-Pérot lines was used along with a hollow-cathode lamp to anchor the numerous spectral features on an absolute scale. We tested our method with the HARPS spectrograph and compared our wavelength solution to the one derived from a laser frequency comb. Results: The combined hollow-cathode lamp/etalon calibration overcomes large distortion (50 m/s) in the wavelength solution of the HARPS data reduction software. The direct comparison to the laser frequency comb shows differences of only 10 m/s at most. Conclusions: Combining hollow-cathode lamps with Fabry-Pérot interferometers can lead to substantial improvements in the wavelength calibration of echelle spectrographs. Etalons can provide economical alternatives to the laser frequency comb, especially for smaller projects.

  6. Calibration of the Herschel SPIRE Fourier Transform Spectrometer

    CERN Document Server

    Swinyard, B M; Hopwood, R; Valtchanov, I; Lu, N; Fulton, T; Benielli, D; Imhof, P; Marchili, N; Baluteau, J -P; Bendo, G J; Ferlet, M; Griffin, M J; Lim, T L; Makiwa, G; Naylor, D A; Orton, G S; Papageorgiou, A; Pearson, C P; Schulz, B; Sidher, S D; Spencer, L D; van der Wiel, M H D; Wu, R

    2014-01-01

    The Herschel SPIRE instrument consists of an imaging photometric camera and an imaging Fourier Transform Spectrometer (FTS), both operating over a frequency range of 450-1550 GHz. In this paper, we briefly review the FTS design, operation, and data reduction, and describe in detail the approach taken to relative calibration (removal of instrument signatures) and absolute calibration against standard astronomical sources. The calibration scheme assumes a spatially extended source and uses the Herschel telescope as primary calibrator. Conversion from extended to point-source calibration is carried out using observations of the planet Uranus. The model of the telescope emission is shown to be accurate to within 6% and repeatable to better than 0.06% and, by comparison with models of Mars and Neptune, the Uranus model is shown to be accurate to within 3%. Multiple observations of a number of point-like sources show that the repeatability of the calibration is better than 1%, if the effects of the satellite absolu...

  7. Stochastic calibration and learning in nonstationary hydroeconomic models

    Science.gov (United States)

    Maneta, M. P.; Howitt, R.

    2014-05-01

    Concern about water scarcity and adverse climate events over agricultural regions has motivated a number of efforts to develop operational integrated hydroeconomic models to guide adaptation and optimal use of water. Once calibrated, these models are used for water management and analysis assuming they remain valid under future conditions. In this paper, we present and demonstrate a methodology that permits the recursive calibration of economic models of agricultural production from noisy but frequently available data. We use a standard economic calibration approach, namely positive mathematical programming, integrated in a data assimilation algorithm based on the ensemble Kalman filter equations to identify the economic model parameters. A moving average kernel ensures that new and past information on agricultural activity are blended during the calibration process, avoiding loss of information and overcalibration for the conditions of a single year. A regularization constraint akin to the standard Tikhonov regularization is included in the filter to ensure its stability even in the presence of parameters with low sensitivity to observations. The results show that the implementation of the PMP methodology within a data assimilation framework based on the enKF equations is an effective method to calibrate models of agricultural production even with noisy information. The recursive nature of the method incorporates new information as an added value to the known previous observations of agricultural activity without the need to store historical information. The robustness of the method opens the door to the use of new remote sensing algorithms for operational water management.

  8. Study of glass hydrometer calibration by hydrostatic weighting

    Science.gov (United States)

    Chen, Chaoyun; Wang, Jintao; Li, Zhihao; Zhang, Peiman

    2016-01-01

    Glass hydrometers are simple but effective instruments for measuring the density of liquids. Glass hydrometers calibration based on the Archimedes law, using silicon ring as a reference standard solid density, n-tridecane with density stability and low surface tension as the standard working liquid, based on hydrostatic weighing method designs a glass hydrometer calibration system. Glass hydrometer calibration system uses CCD image measurement system to align the scale of hydrometer and liquid surface, with positioning accuracy of 0.01 mm. Surface tension of the working liquid is measured by Whihemy plate. According to twice glass hydrometer weighing in the air and liquid can calculate the correction value of the current scale. In order to verify the validity of the principle of the hydrostatic weighing method of glass hydrometer calibration system, for measuring the density range of (770-790) kg/m3, with a resolution of 0.2 kg/m3 of hydrometer. The results of measurement compare with the Physikalisch-Technische Bundesanstalt(PTB) ,verifying the validity of the calibration system.

  9. Review of Scattering Corrections for Calibration of Neutron Survey Meters

    International Nuclear Information System (INIS)

    In this study, the calibration factors of the several neutron survey meters obtained by both the shadow-cone method and semi-empirical method. Generally and theoretically, the calibration factors obtained by the methods given in ISO 8529-2 have the similar values. However we found that the calibration factors obtained by two methods have the different values. The calibration factors were obtained by two calibration method and the validity of the 'shadow-cone method' was proven by the MCNPX calculation. The total scattering correction of the conventional true scatter effects in the 'shadow-cone method' was under-estimated by 4.0.8.1% rather than the scatter correction in the 'semi-empirical method'. This lower calibration factor than that of the 'semi-empirical method' is owing to the under-response of the neutron survey meters in the thermal neutron field. These results are elucidated from that most of single-moderator-based survey meters excluding tissue-equivalent proportional counter (TEPC) have an under-response in the thermal neutron field. It is concluded that a great care is needed while calibrating a survey meter using a shadow-cone calibration method and in interpreting the readouts. ISO 8529-2 (the International Organization for Standardization) recommends four different approaches to the problem of correcting for scatter effects. The first three methods, denoted as the shadow-cone method, the generalized fit method, and the semi-empirical method, usually involve an initial set of careful measurements as a function of the distance between neutron source and detector

  10. A Method to Test Model Calibration Techniques: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, Ron; Polly, Ben; Neymark, Joel

    2016-09-01

    This paper describes a method for testing model calibration techniques. Calibration is commonly used in conjunction with energy retrofit audit models. An audit is conducted to gather information about the building needed to assemble an input file for a building energy modeling tool. A calibration technique is used to reconcile model predictions with utility data, and then the 'calibrated model' is used to predict energy savings from a variety of retrofit measures and combinations thereof. Current standards and guidelines such as BPI-2400 and ASHRAE-14 set criteria for 'goodness of fit' and assume that if the criteria are met, then the calibration technique is acceptable. While it is logical to use the actual performance data of the building to tune the model, it is not certain that a good fit will result in a model that better predicts post-retrofit energy savings. Therefore, the basic idea here is that the simulation program (intended for use with the calibration technique) is used to generate surrogate utility bill data and retrofit energy savings data against which the calibration technique can be tested. This provides three figures of merit for testing a calibration technique, 1) accuracy of the post-retrofit energy savings prediction, 2) closure on the 'true' input parameter values, and 3) goodness of fit to the utility bill data. The paper will also discuss the pros and cons of using this synthetic surrogate data approach versus trying to use real data sets of actual buildings.

  11. Fusion yield measurements on JET and their calibration

    Energy Technology Data Exchange (ETDEWEB)

    Syme, D.B., E-mail: brian.syme@ccfe.ac.uk [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Popovichev, S. [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Conroy, S. [EURATOM-VR Association, Department of Physics and Astronomy, Uppsala University, Box 516, SE-75120 Uppsala (Sweden); Lengar, I.; Snoj, L. [EURATOM-MHEST Association, Reactor Physics Division, Jožef Stefan Institute, Jamova cesta 39, SI-1000 Ljubljana (Slovenia); Sowden, C. [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Giacomelli, L. [EURATOM-ENEA-CNR Association, CNR-IFP and Univ. di Milano-Bicocca, Milan (Italy); Hermon, G.; Allan, P.; Macheta, P.; Plummer, D.; Stephens, J. [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Batistoni, P. [EURATOM-ENEA Association, Via E. Fermi,40, 00044 Frascati (Italy); Prokopowicz, R.; Jednorog, S. [EURATOM-IPPLM Association, Institute of Plasma Physics and Laser Microfusion, Hery 23, 01-497 Warsaw (Poland); Abhangi, M.R.; Makwana, R. [Institute for Plasma Research, Bhat, Gandhinagar, 382 428 Gujarat (India)

    2014-11-15

    The power output of fusion experiments and fusion reactor-like devices is measured in terms of the neutron yields which relate directly to the fusion yield. In this paper we describe the devices and methods used to make the new in situ calibration of JET in April 2013 and its early results. The target accuracy of this calibration was 10%, just as in the earlier JET calibration and as required for ITER, where a precise neutron yield measurement is important, e.g., for tritium accountancy. We discuss the constraints and early decisions which defined the main calibration approach, e.g., the choice of source type and the deployment method. We describe the physics, source issues, safety and engineering aspects required to calibrate directly the Fission Chambers and the Activation System which carry the JET neutron calibration. In particular a direct calibration of the Activation system was planned for the first time in JET. We used the existing JET remote-handling system to deploy the {sup 252}Cf source and developed the compatible tooling and systems necessary to ensure safe and efficient deployment in these cases. The scientific programme has sought to better understand the limitations of the calibration, to optimise the measurements and other provisions, to provide corrections for perturbing factors (e.g., presence of the remote-handling boom and other non-standard torus conditions) and to ensure personnel safety and safe working conditions. Much of this work has been based on an extensive programme of Monte-Carlo calculations which, e.g., revealed a potential contribution to the neutron yield via a direct line of sight through the ports which presents individually depending on the details of the port geometry.

  12. Hierarchical Bayesian Data Analysis in Radiometric SAR System Calibration: A Case Study on Transponder Calibration with RADARSAT-2 Data

    Directory of Open Access Journals (Sweden)

    Björn J. Döring

    2013-12-01

    Full Text Available A synthetic aperture radar (SAR system requires external absolute calibration so that radiometric measurements can be exploited in numerous scientific and commercial applications. Besides estimating a calibration factor, metrological standards also demand the derivation of a respective calibration uncertainty. This uncertainty is currently not systematically determined. Here for the first time it is proposed to use hierarchical modeling and Bayesian statistics as a consistent method for handling and analyzing the hierarchical data typically acquired during external calibration campaigns. Through the use of Markov chain Monte Carlo simulations, a joint posterior probability can be conveniently derived from measurement data despite the necessary grouping of data samples. The applicability of the method is demonstrated through a case study: The radar reflectivity of DLR’s new C-band Kalibri transponder is derived through a series of RADARSAT-2 acquisitions and a comparison with reference point targets (corner reflectors. The systematic derivation of calibration uncertainties is seen as an important step toward traceable radiometric calibration of synthetic aperture radars.

  13. Comprehensive Calibration and Validation Site for Information Remote Sensing

    Science.gov (United States)

    Li, C. R.; Tang, L. L.; Ma, L. L.; Zhou, Y. S.; Gao, C. X.; Wang, N.; Li, X. H.; Wang, X. H.; Zhu, X. H.

    2015-04-01

    As a naturally part of information technology, Remote Sensing (RS) is strongly required to provide very precise and accurate information product to serve industry, academy and the public at this information economic era. To meet the needs of high quality RS product, building a fully functional and advanced calibration system, including measuring instruments, measuring approaches and target site become extremely important. Supported by MOST of China via national plan, great progress has been made to construct a comprehensive calibration and validation (Cal&Val) site, which integrates most functions of RS sensor aviation testing, EO satellite on-orbit caration and performance assessment and RS product validation at this site located in Baotou, 600km west of Beijing. The site is equipped with various artificial standard targets, including portable and permanent targets, which supports for long-term calibration and validation. A number of fine-designed ground measuring instruments and airborne standard sensors are developed for realizing high-accuracy stepwise validation, an approach in avoiding or reducing uncertainties caused from nonsynchronized measurement. As part of contribution to worldwide Cal&Val study coordinated by CEOS-WGCV, Baotou site is offering its support to Radiometric Calibration Network of Automated Instruments (RadCalNet), with an aim of providing demonstrated global standard automated radiometric calibration service in cooperation with ESA, NASA, CNES and NPL. Furthermore, several Cal&Val campaigns have been performed during the past years to calibrate and validate the spaceborne/airborne optical and SAR sensors, and the results of some typical demonstration are discussed in this study.

  14. Comparative study among calibration methods of clinical applicators of beta radiation

    International Nuclear Information System (INIS)

    90Sr+90Y clinical applicators are instruments used in brachytherapy procedures and they have to be periodically calibrated, according to international standards and recommendations. In this work, four calibration methods of dermatological and ophthalmic applicators were studied, comparing the results with those given by the calibration certificates of the manufacturers. The methods included the use of the standard applicator of the Calibration Laboratory (LCI), calibrated by the National Institute of Standards and Technology; an Amersham applicator (LCI) as reference; a mini-extrapolation chamber developed at LCI as an absolute standard; and thermoluminescent dosimetry. The mini-extrapolation chamber and a PTW commercial extrapolation chamber were studied in relation to their performance through quality control tests of their response, as leakage current, repeatability and reproducibility. The distribution of the depth dose in water, that presents high importance in dosimetry of clinical applicators, was determined using the mini extrapolation chamber and the thermoluminescent dosimeters. The results obtained were considered satisfactory for the both cases, and comparable to the data of the IAEA (2002) standard. Furthermore, a dosimetry postal kit was developed for the calibration of clinical applicators using the thermoluminescent technique, to be sent to clinics and hospitals, without the need of the transport of the sources to IPEN for calibration. (author)

  15. A least squares procedure for calculating the calibration constants of a portable gamma-ray spectrometer.

    Science.gov (United States)

    Ribeiro, F B; Carlos, D U; Hiodo, F Y; Strobino, E F

    2005-01-01

    In this study, a least squares procedure for calculating the calibration constants of a portable gamma-ray spectrometer using the general inverse matrix method is presented. The procedure weights the model equations fitting to the calibration data, taking into account the variances in the counting rates and in the radioactive standard concentrations. The application of the described procedure is illustrated by calibrating twice the same gamma-ray spectrometer, with two independent data sets collected approximately 18 months apart in the same calibration facility.

  16. Liquid scintillation counting as a calibration method for PAEC monitors

    International Nuclear Information System (INIS)

    Measurements of potential alpha energy concentration (PAEC) of radon progeny are obligatory in Polish coal mines. Very important problem is a proper calibration of PAEC monitors for coal mining industry. It is very know fact, that is no possibility of producing of standard atmosphere of radon progeny. Therefore calibration is usually based on comparison of read-out of PAEC monitor with more counting of this filter immersed in a liquid scintillation counter (LSC) gives as a result an absolute method of PAEC measurements, because of very high efficiency for alpha and beta particles in liquid scintillator (almost 100%). (author). 30 refs, 2 figs

  17. Calibration of Frequency Data Collection Systems Using Shortwave Radio Signals

    Science.gov (United States)

    Estler, Ron

    2000-09-01

    The atomic-clock-derived audio tones broadcast on the National Institute of Standards and Technology (NIST) shortwave station WWV are used to calibrate computer frequency data collection systems via Fast Fourier Transforms (FFT). Once calibrated, the data collection system can be used to accurately determine the audio signals used in several instructional physical chemistry laboratory experiments. This method can be applied to virtually any hardware-software configuration that allows adjustment of the apparent time scale (digitizing rate) of the recorded audio file.

  18. A pratical method of data elimination in radiometric calibration

    International Nuclear Information System (INIS)

    The absence of phosphorite standards to uranium radiometric analyses sugests the necessity of intercomparison samples to construct an uranium calibration curve. The intercomparison analysis was performed by delay neutrons, X-rays fluorescence and molecular absorption spectrophotometry in three diferents laboratories. This work intends to stablish rejection criteria for deviating data in order to achieve better precision. Through the comparison from results of the analysis methods, the data outside the confidence intervals, were eliminated. A more accurate calibration curve inside the interval +-1,1% to +-3.0% was obtained with this method. (author)

  19. Generic calibration procedures for nacelle-based profiling lidars

    DEFF Research Database (Denmark)

    Borraccino, Antoine; Courtney, Michael; Wagner, Rozenn

    In power performance testing, it has been demonstrated that the effects of wind speed and direction variations over the rotor disk can no longer be neglected for large wind turbines [1]. A new generation of commercial nacelle-based lidars is now available, offering wind profiling capabilities....... Developing standard procedures for power curves using lidars requires assessing lidars measurement uncertainty that is provided by a calibration. Based on the calibration results from two lidars, the Avent 5-beam Demonstrator and the Zephir Dual Mode (ZDM), we present in this paper a generic methodology...

  20. The KLOE Online Calibration System

    Institute of Scientific and Technical Information of China (English)

    E.Pasqualucci

    2001-01-01

    Based on all the features of the KLOE online software,the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed.Acalibration manager process controls the system,implementing the interface to the online system,receiving information from the run control and translating its state transitions to a separate state machine.It acts as a " calibration run controller"and performs failure recovery when requested by a set of process checkers.The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms.A client library and C,fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an bool-like interface.Several calibration processes running in parallel in a destributed,multiplatform environment can fill the same histograms,allowing fast external information check.A monitor thread allow remote browsing for visual inspection,Pre-filtered data are read in nonprivileged spy mode from the data acquisition system via the Kloe Integrated Dataflow,privileged spy mode from the data acquisiton system via the Kole Integrated Dataflow.The main characteristics of the system are presented.

  1. Court Reconstruction for Camera Calibration in Broadcast Basketball Videos.

    Science.gov (United States)

    Wen, Pei-Chih; Cheng, Wei-Chih; Wang, Yu-Shuen; Chu, Hung-Kuo; Tang, Nick C; Liao, Hong-Yuan Mark

    2016-05-01

    We introduce a technique of calibrating camera motions in basketball videos. Our method particularly transforms player positions to standard basketball court coordinates and enables applications such as tactical analysis and semantic basketball video retrieval. To achieve a robust calibration, we reconstruct the panoramic basketball court from a video, followed by warping the panoramic court to a standard one. As opposed to previous approaches, which individually detect the court lines and corners of each video frame, our technique considers all video frames simultaneously to achieve calibration; hence, it is robust to illumination changes and player occlusions. To demonstrate the feasibility of our technique, we present a stroke-based system that allows users to retrieve basketball videos. Our system tracks player trajectories from broadcast basketball videos. It then rectifies the trajectories to a standard basketball court by using our camera calibration method. Consequently, users can apply stroke queries to indicate how the players move in gameplay during retrieval. The main advantage of this interface is an explicit query of basketball videos so that unwanted outcomes can be prevented. We show the results in Figs. 1, 7, 9, 10 and our accompanying video to exhibit the feasibility of our technique. PMID:27504515

  2. Court Reconstruction for Camera Calibration in Broadcast Basketball Videos.

    Science.gov (United States)

    Wen, Pei-Chih; Cheng, Wei-Chih; Wang, Yu-Shuen; Chu, Hung-Kuo; Tang, Nick C; Liao, Hong-Yuan Mark

    2016-05-01

    We introduce a technique of calibrating camera motions in basketball videos. Our method particularly transforms player positions to standard basketball court coordinates and enables applications such as tactical analysis and semantic basketball video retrieval. To achieve a robust calibration, we reconstruct the panoramic basketball court from a video, followed by warping the panoramic court to a standard one. As opposed to previous approaches, which individually detect the court lines and corners of each video frame, our technique considers all video frames simultaneously to achieve calibration; hence, it is robust to illumination changes and player occlusions. To demonstrate the feasibility of our technique, we present a stroke-based system that allows users to retrieve basketball videos. Our system tracks player trajectories from broadcast basketball videos. It then rectifies the trajectories to a standard basketball court by using our camera calibration method. Consequently, users can apply stroke queries to indicate how the players move in gameplay during retrieval. The main advantage of this interface is an explicit query of basketball videos so that unwanted outcomes can be prevented. We show the results in Figs. 1, 7, 9, 10 and our accompanying video to exhibit the feasibility of our technique.

  3. Multi-residue pesticide analysis (gas chromatography-tandem mass spectrometry detection)-Improvement of the quick, easy, cheap, effective, rugged, and safe method for dried fruits and fat-rich cereals-Benefit and limit of a standardized apple purée calibration (screening).

    Science.gov (United States)

    Rasche, Claudia; Fournes, Britta; Dirks, Uwe; Speer, Karl

    2015-07-17

    Some steps of the QuEChERS method for the analysis of pesticides with GC-MS/MS in cereals and dried fruits were improved or simplified. For the latter, a mixing vessel with stator-rotor-system proved to be advantageous. The extraction procedure of dried fruits is much easier and safer than the Ultra Turrax and results in excellent validation data at a concentration level of 0.01mg/kg (116 of 118 analytes with recoveries in the range of 70-120%, 117 of 118 analytes with RSD 7%), predominantly organochlorines showed recoveries of extraction was carried out analogous to the QuEChERS method, however, without the addition of water. With this simple modification, the problematic lipophilic pesticides, which had been strongly affected by the fat content of the commodities, could be determined with recoveries above 70% even at a concentration level of 0.01mg/kg. Moreover, a GC-MS/MS screening method for 120 pesticides at a concentration level of 0.01mg/kg was established by employing analyte protectants (ethylglycerol, gulonolactone, and sorbitol). The use of only one standardized calibration, made of an apple purée extract in combination with analyte protectants, allowed for a qualitative and quantitative analysis of 120 pesticides in different matrix extracts (tomato, red pepper, sour cherries, dried apples, black currant powder, raisins, wheat flour, rolled oats, wheat germ). The analyte protectants leveled the differences in the matrix-induced protection effect of the analyzed extracts over a wide range. The majority of the pesticides were analyzed with good analytical results (recoveries in the range of 70-120% and RSD <20%). PMID:26044382

  4. DOE radiological calibrations intercomparison program: Results of fiscal year 1986

    International Nuclear Information System (INIS)

    The Department of Energy Radiological Calibration Intercomparison Program was initiated in January 1986, under the research portion of the DOE Laboratory Accreditation Program. The program operates via the exchange of transfer standards, consisting of instrument sets and standard secondary beta sources. There are two instrument sets and the scheduled use has been staggered such that one set is available for use during each month. One set of secondary standard beta sources is available for use bimonthly. During the 1986 fiscal year, five laboratories used the instrument sets and three laboratories used the beta source set. Results were reported for all the measurements. The average and one standard deviation of the ratios of participant results to Pacific Northwest Laboratory calibration values were 1.12 +- 0.17 for gamma measurements. Those ratios for the gamma measurements varied from 0.98 to 3.06. The larger differences of results from measurements performed at two facilities were directly attributable to unfamiliarity with the intercomparison instruments. The average and one standard deviation of the ratios of participant results to PNL calibration values obtained using the secondary 90Sr beta source was 1.02 +- 0.05, which is well within measurement uncertainties. The one participant who performed measurements using 147Pm and 204Tl sources obtained ratios of 0.68 and 1.11, respectively. No measurements were performed using neutron or x-ray sources

  5. DOE radiological calibrations intercomparison program: Results of fiscal year 1986

    Energy Technology Data Exchange (ETDEWEB)

    Cummings, F.M.; Roberson, P.L.; McDonald, J.C.

    1987-05-01

    The Department of Energy Radiological Calibration Intercomparison Program was initiated in January 1986, under the research portion of the DOE Laboratory Accreditation Program. The program operates via the exchange of transfer standards, consisting of instrument sets and standard secondary beta sources. There are two instrument sets and the scheduled use has been staggered such that one set is available for use during each month. One set of secondary standard beta sources is available for use bimonthly. During the 1986 fiscal year, five laboratories used the instrument sets and three laboratories used the beta source set. Results were reported for all the measurements. The average and one standard deviation of the ratios of participant results to Pacific Northwest Laboratory calibration values were 1.12 +- 0.17 for gamma measurements. Those ratios for the gamma measurements varied from 0.98 to 3.06. The larger differences of results from measurements performed at two facilities were directly attributable to unfamiliarity with the intercomparison instruments. The average and one standard deviation of the ratios of participant results to PNL calibration values obtained using the secondary /sup 90/Sr beta source was 1.02 +- 0.05, which is well within measurement uncertainties. The one participant who performed measurements using /sup 147/Pm and /sup 204/Tl sources obtained ratios of 0.68 and 1.11, respectively. No measurements were performed using neutron or x-ray sources.

  6. Development of in-situ calibration method's algorithm for thermal imager

    OpenAIRE

    Dzikovska, Yu.; Hots, N.

    2015-01-01

    The effect of impact factors on the output signal of the radiation receiver is analyzed. Necessity to do additional in-situ calibration of thermal imager by using the standard extended area grey emitter is justified. In-situ calibration method's algorithm for thermal imager is developed. It helps to eliminate the effect of impact factors on the measurement results of temperature gradient. The requirements to the standard extended area emitter are formulated.

  7. Power calibrations for TRIGA reactors

    International Nuclear Information System (INIS)

    The purpose of this paper is to establish a framework for the calorimetric power calibration of TRIGA reactors so that reliable results can be obtained with a precision better than ± 5%. Careful application of the same procedures has produced power calibration results that have been reproducible to ± 1.5%. The procedures are equally applicable to the Mark I, Mark II and Mark III reactors as well as to reactors having much larger reactor tanks and to TRIGA reactors capable of forced cooling up to 3 MW in some cases and 15 MW in another case. In the case of forced cooled TRIGA reactors, the calorimetric power calibration is applicable in the natural convection mode for these reactors using exactly the same procedures as are discussed below for the smaller TRIGA reactors (< 2 MW)

  8. Facility for dosimetric instrument calibration

    International Nuclear Information System (INIS)

    A structure is designed consisting of a rotary support of containers with radiation sources and of a rotary plug mounted above the sources. A support post with a slide rest and arms rotating around the post longitudinal axis is mounted in the centre of the container support. THe arms support instruments to be calibrated. The colimation cone of the respective source is directed to the tensor of the instrument being calibrated. The slide rest is balanced using a counterpoise mounted in the support post. Sources are not removed from the containers in source change during measurement. The support can hold more containers and the slide rest can support more instruments to be calibrated than the existing configurations. (M.D.). 2 figs

  9. Evaluating uncertainties in the cross-calibration of parallel ion chambers used in electron beam radiotherapy; Avaliacao das incertezas no processo de calibracao cruzada de camaras de placas paralelas utilizadas em feixes de eletrons para radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Anderson, Ernani; Travassos, Paulo; Ferreira, Max da Silva; Carvalho, Samira Marques de; Silva, Michele Maria da; Peixoto, Jose Guilherme Pereira, E-mail: ernanianderson@hotmail.com [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil); Salmon Junior, Helio Augusto [Grupo COI - Clinicas Oncologicas, Rio de Janeiro, RJ (Brazil)

    2015-07-01

    This study aims to estimative the combined standard uncertainty for a detector parallel plate used for dosimetry of electron beams in linear accelerators for radiotherapy, which has been calibrated by the cross-calibration method. Keeping the combined standard uncertainty next of the uncertainty informed in the calibration certificate of the reference chamber, become possible establish the calibration factor of the detector. The combined standard uncertainty obtained in this study was 2.5 %. (author)

  10. Calibration Report for the WRAP Facility Gamma Energy Analysis System [104-ND-06-102A

    Energy Technology Data Exchange (ETDEWEB)

    WILLS, C.E.

    2000-02-15

    The Waste Receiving And Processing facility (WRAP) adheres to providing gamma-ray spectroscopy instrument calibrations traceable to the National Institute for Standards and Technology (NIST) standards. The detectors are used to produce quantitative results for the Waste Isolation Pilot Plant (WIPP) and must meet calibration programmatic calibration goals. Instruments must meet portions of ANSI N42.14, 1978 guide for Germanium detectors. The Non-Destructive Assay (NDA) Gamma Energy Analysis (GEA) utilizes NIST traceable line source standards for the detector system calibrations. The counting configuration is a series of drums containing the line sources and different density filler matrices. The drums are used to develop system efficiencies with respect to density. The efficiency and density correction factors are required for the processing of drummed waste materials of similar densities. The calibration verification is carried out after the calibration is deemed final, by counting a second drum of NIST traceable sources. Three in-depth calibrations have been completed on one of the two systems to date, the first being the system acceptance plan. This report has a secondary function; that being the development of the instrument calibration errors which are to be folded into the Total Instrument Uncertainty document, HNF-4050.

  11. Calibration Report for the WRAP Facility Gamma Energy Analysis System (104-ND-06-102A)

    Energy Technology Data Exchange (ETDEWEB)

    WILLS, C.E.

    2000-03-13

    The Waste Receiving And Processing facility (WRAP) adheres to providing gamma-ray spectroscopy instrument calibrations traceable to the National Institute for Standards and Technology (NIST) standard{sup (4)}. The detectors are used to produce quantitative results for the Waste Isolation Pilot Plant (WIPP) and must meet calibration programmatic calibration goals. Instruments must meet portions of ANSI N42.14, 1978 guide for Germanium detectors. The Non-Destructive Assay (NDA) Gamma Energy Analysis (GEA) utilizes NIST traceable line source standards for the detector system calibrations. The counting configuration is a series of drums containing the line sources and different density filler matrices. The drums are used to develop system efficiencies with respect to density. The efficiency and density correction factors are required for the processing of drummed waste materials of similar densities. The calibration verification is carried out after the calibration is deemed final, by counting a second drum of NIST traceable sources. Three in-depth calibrations have been completed on one of the two systems to date, the first being the system acceptance plan. This report has a secondary function; that being the development of the instrument calibration errors which are to be folded into the Total Instrument Uncertainty document, HNF-4050.

  12. Quantitative Analysis of Sulfide Minerals by Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry Using Glass Reference Materials with Matrix Normalization Plus Sulfur Internal Standardization Calibration%玻璃标样结合硫内标归一定量技术在激光剥蚀-等离子体质谱分析硫化物矿物中的应用

    Institute of Scientific and Technical Information of China (English)

    袁继海; 詹秀春; 范晨子; 赵令浩; 孙冬阳; 贾泽荣; 胡明月; 蒯丽君

    2012-01-01

    A novel strategy for microanalysis of sulfides by laser ablation (LA)-ICP-MS was established. In this method, the relative sensitivity factor of Ca relative to S in anhydrite mineral reference material was taken as a transition bridge, by which the relative sensitivity factors of interesting elements in glass reference materials relative to Ca could be converted into relative sensitivity factors relative to S by the transition bridge, then the quantitative analysis of multi-elements in sulfide minerals by multi-glass reference materials with matrix normalization plus sulfur internal standardizationcalibration was carried out. 20 elements in the American polymetal sulfide mineral reference material MASS-1 were analyzed using this new method. The relative errors of major elements in MASS-1 were less than 10% ? And the results of trace elements with reference values were nearly within the uncertainty of the preliminary values. Multi-elements in 12 sulfide single minerals were analyzed by applying this new method. The relative errors of the greatest number of major elements were less than 10% , with which the results of most major elements were accurate than those obtained by MASS-1 as calibration standard with matrix normalization plus internal standardization or internal standard calibration. And the results to trace elements agreed well with the calibrated results by MASS-1 with matrix normalization plus internal standardization or internal standard calibration. This method overcomes the problem of non-matrix matched standards, enables to accurately determine the major composition of sulfur in sulfide minerals and suggests a new approach for analysis of sulfide minerals.%以硬石膏矿物标样中Ca相对于S的灵敏度因子为基准,将玻璃标样中主量和痕量元素相对于Ca的灵敏度因子转换成元素相对于S的灵敏度因子,建立了多玻璃标样结合硫内标归一定量技术分析硫化物单矿物多元素的新方法.利用本方法

  13. 1 {\\Omega}-10 k{\\Omega} high precision transportable setup to calibrate multifunction electrical instruments

    CERN Document Server

    Capra, P P

    2015-01-01

    A temperature controlled 1 {\\Omega}-10 k{\\Omega} standard Resistors transportable setup was developed at National Institute of Metrological Research, (INRIM) for the calibration and adjustment of multifunction electrical instruments. The two Standards consist respectively of two 10 {\\Omega} and 100 k{\\Omega} parallel connected resistors nets inserted in a temperature controlled aluminium box. Novelty of the realization is the oil insertion of the 1 {\\Omega} net with its internal connectors lowering the thermo-electromotive forces (emfs) effects. Short and mid-term stabilities of the setup Standards resulted on the order and in some cases better than other top level 1 {\\Omega} and 10 k{\\Omega} commercial Standards. The transport effect turning off the setup temperature control did not cause appreciable measurement deviations on the two Standards. The Standards uncertainties meet those requested by DMMs and MFCs manufacturers to calibrate and adjust these instruments. A test to adjust a multifunction calibrator...

  14. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    Science.gov (United States)

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

  15. Calibrating Data from the Hinode/X-Ray Telescope and Associated Uncertainties

    CERN Document Server

    Kobelski, Adam R; Weber, Mark A; McKenzie, David E; Reeves, Katharine K

    2013-01-01

    The X-Ray Telescope (XRT) onboard the Hinode satellite, launched 23 September 2006 by the Japanese Aerospace Exploration Agency (JAXA) is a joint mission between Japan, the United States, and the United Kingdom to study the solar corona. In particular XRT was designed to study solar plasmas with temperatures between 1 and 10 MK with $\\approx1''$ pixels ($\\approx2''$ resolution). Prior to analysis, the data product from this instrument must be properly calibrated and data values quantified in order to assess accurately the information contained within. We present here the standard methods of calibration for these data. The calibration is performed on an empirical basis which uses the least complicated correction that accurately describes the data while suppressing spurious features. By analyzing the uncertainties remaining in the data after calibration, we conclude that the procedure is successful, as the remaining uncertainty after calibration is dominated by photon noise. This calibration software is availab...

  16. NASA's Metrology and Calibration program

    Science.gov (United States)

    Kern, Frederick A.

    1991-01-01

    The success of the National Aeronautics and Space Administration's scientific and engineering projects is often based on the capability and quality of the metrology and calibration programs of its field centers and facilities. A Metrology and Calibration Working Group, having representation from each of the NASA field centers and facilities, was established to develop and implement a program to provide a centralized direction and agency wide focus. A brief history of the Group is presented. The development of the Group's operating philosophy, the long term objectives, the measurement research and development program, and review of accomplishments is discussed.

  17. GREAT/SOFIA atmospheric calibration

    CERN Document Server

    Guan, Xin; Graf, Urs U; Güsten, Rolf; Okada, Yoko; Torres, Miguel Angel Requena; Simon, Robert; Wiesemeyer, Helmut

    2012-01-01

    The GREAT observations need frequency-selective calibration across the passband for the residual atmospheric opacity at flight altitude. At these altitudes the atmospheric opacity has both narrow and broad spectral features. To determine the atmospheric transmission at high spectral resolution, GREAT compares the observed atmospheric emission with atmospheric model predictions, and therefore depends on the validity of the atmospheric models. We discusse the problems identified in this comparison with respect to the observed data and the models, and describe the strategy used to calibrate the science data from GREAT/SOFIA during the first observing periods.

  18. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Kristensen, L.; Jensen, G.; Hansen, A.;

    2001-01-01

    . It is concluded that the method has the advantage that many anemometers can be calibrated accurately with a minimum of work and cost. The obvious disadvantage is that the calibration of a set of anemometersmay take more than one month in order to have wind speeds covering a sufficiently large magnitude range...... in a wind direction sector where we can be sure that the instruments are exposed to identical, simultaneous wind flows. Another main conclusion is thatstatistical uncertainty must be carefully evaluated since the individual 10 minute wind-speed averages are not statistically independent....

  19. Effective radiation attenuation calibration for breast density: compression thickness influences and correction

    Directory of Open Access Journals (Sweden)

    Thomas Jerry A

    2010-11-01

    Full Text Available Abstract Background Calibrating mammograms to produce a standardized breast density measurement for breast cancer risk analysis requires an accurate spatial measure of the compressed breast thickness. Thickness inaccuracies due to the nominal system readout value and compression paddle orientation induce unacceptable errors in the calibration. Method A thickness correction was developed and evaluated using a fully specified two-component surrogate breast model. A previously developed calibration approach based on effective radiation attenuation coefficient measurements was used in the analysis. Water and oil were used to construct phantoms to replicate the deformable properties of the breast. Phantoms consisting of measured proportions of water and oil were used to estimate calibration errors without correction, evaluate the thickness correction, and investigate the reproducibility of the various calibration representations under compression thickness variations. Results The average thickness uncertainty due to compression paddle warp was characterized to within 0.5 mm. The relative calibration error was reduced to 7% from 48-68% with the correction. The normalized effective radiation attenuation coefficient (planar representation was reproducible under intra-sample compression thickness variations compared with calibrated volume measures. Conclusion Incorporating this thickness correction into the rigid breast tissue equivalent calibration method should improve the calibration accuracy of mammograms for risk assessments using the reproducible planar calibration measure.

  20. Self-calibrating lateral scanning white-light interferometer

    Energy Technology Data Exchange (ETDEWEB)

    Munteanu, Florin

    2010-04-20

    Lateral scanning white-light interferometry represents an attractive alternative to the standard white-light interferometry. Its main advantage over the latter procedure consists in the ability to scan large samples continuously, without the need of a cumbersome stitching procedure. Presently, the main drawback in the path of large-scale industrial acceptance of this method is the need for careful calibration of the tilt angle prior to each measurement. A novel self-calibration approach is presented. Using the data acquired during the normal scanning process, the need of an initial tilt angle calibration is eliminated and on-the-fly system adjustments for the best signal-to-noise ratio can be performed without an increase in the measurement time dictated by recalibration.

  1. A Generic Algorithm for IACT Optical Efficiency Calibration using Muons

    CERN Document Server

    Mitchell, A M W; Parsons, R D

    2015-01-01

    Muons produced in Extensive Air Showers (EAS) generate ring-like images in Imaging Atmospheric Cherenkov Telescopes when travelling near parallel to the optical axis. From geometrical parameters of these images, the absolute amount of light emitted may be calculated analytically. Comparing the amount of light recorded in these images to expectation is a well established technique for telescope optical efficiency calibration. However, this calculation is usually performed under the assumption of an approximately circular telescope mirror. The H.E.S.S. experiment entered its second phase in 2012, with the addition of a fifth telescope with a non-circular 600m$^2$ mirror. Due to the differing mirror shape of this telescope to the original four H.E.S.S. telescopes, adaptations to the standard muon calibration were required. We present a generalised muon calibration procedure, adaptable to telescopes of differing shapes and sizes, and demonstrate its performance on the H.E.S.S. II array.

  2. Calibration technology in application of robot-laser scanning system

    Science.gov (United States)

    Ren, YongJie; Yin, ShiBin; Zhu, JiGui

    2012-11-01

    A system composed of laser sensor and 6-DOF industrial robot is proposed to obtain complete three-dimensional (3-D) information of the object surface. Suitable for the different combining ways of laser sensor and robot, a new method to calibrate the position and pose between sensor and robot is presented. By using a standard sphere with known radius as a reference tool, the rotation and translation matrices between the laser sensor and robot are computed, respectively in two steps, so that many unstable factors introduced in conventional optimization methods can be avoided. The experimental results show that the accuracy of the proposed calibration method can be achieved up to 0.062 mm. The calibration method is also implemented into the automated robot scanning system to reconstruct a car door panel.

  3. The New Approach to Camera Calibration - GCPs or TLS Data?

    Science.gov (United States)

    Markiewicz, J.; Podlasiak, P.; Kowalczyk, M.; Zawieska, D.

    2016-06-01

    Camera calibration is one of the basic photogrammetric tasks responsible for the quality of processed products. The majority of calibration is performed with a specially designed test field or during the self-calibration process. The research presented in this paper aims to answer the question of whether it is necessary to use control points designed in the standard way for determination of camera interior orientation parameters. Data from close-range laser scanning can be used as an alternative. The experiments shown in this work demonstrate the potential of laser measurements, since the number of points that may be involved in the calculation is much larger than that of commonly used ground control points. The problem which still exists is the correct and automatic identification of object details in the image, taken with a tested camera, as well as in the data set registered with the laser scanner.

  4. Providing radiometric traceability for the calibration home base of DLR by PTB

    Energy Technology Data Exchange (ETDEWEB)

    Taubert, D. R.; Hollandt, J.; Sperfeld, P.; Pape, S.; Hoepe, A.; Hauer, K.-O. [Physikalisch-Technische Bundesanstalt, Braunschweig und Berlin, 10587 Berlin (Germany); Gege, P.; Schwarzmaier, T.; Lenhard, K.; Baumgartner, A. [Deutsches Zentrum fuer Luft- und Raumfahrt, Institut fuer Methodik der Fernerkundung, 82234 Oberpfaffenhofen (Germany)

    2013-05-10

    A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance E{sub {lambda}}({lambda}) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor {beta}{sub 0 Degree-Sign :45 Degree-Sign }({lambda}) of the RASTA reflection standard was calibrated in a 0 Degree-Sign :45 Degree-Sign -viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k= 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independent from procedure I and allowed to cover the entire spectral range of RASTA from 350 nm to 2500 nm. In the second procedure, the 0 Degree-Sign :45 Degree-Sign -viewing geometry spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign }({lambda}) of RASTA was directly calibrated at the Spectral Radiance Comparator Facility (SRCF) of PTB. The relative uncertainties for this calibration procedure range from 0.8 % in the visible up to 7.5 % at 2500 nm (k= 1). In the overlapping spectral range of both calibration procedures the calculated spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign ,calc}({lambda}) from procedure I is in good agreement with the direct measurement of procedure II, i.e. well within the combined expanded uncertainties (k= 2) of both procedures.

  5. Wind Tunnel Balance Calibration: Are 1,000,000 Data Points Enough?

    Science.gov (United States)

    Rhew, Ray D.; Parker, Peter A.

    2016-01-01

    Measurement systems are typically calibrated based on standard practices established by a metrology standards laboratory, for example the National Institute for Standards and Technology (NIST), or dictated by an organization's metrology manual. Therefore, the calibration is designed and executed according to an established procedure. However, for many aerodynamic research measurement systems a universally accepted standard, traceable approach does not exist. Therefore, a strategy for how to develop a calibration protocol is left to the developer or user to define based on experience and recommended practice in their respective industry. Wind tunnel balances are one such measurement system. Many different calibration systems, load schedules and procedures have been developed for balances with little consensus on a recommended approach. Especially lacking is guidance the number of calibration data points needed. Regrettably, the number of data points tends to be correlated with the perceived quality of the calibration. Often, the number of data points is associated with ones ability to generate the data rather than by a defined need in support of measurement objectives. Hence the title of the paper was conceived to challenge recent observations in the wind tunnel balance community that shows an ever increasing desire for more data points per calibration absent of guidance to determine when there are enough. This paper presents fundamental concepts and theory to aid in the development of calibration procedures for wind tunnel balances and provides a framework that is generally applicable to the characterization and calibration of other measurement systems. Questions that need to be answered are for example: What constitutes an adequate calibration? How much data are needed in the calibration? How good is the calibration? This paper will assist a practitioner in answering these questions by presenting an underlying theory on how to evaluate a calibration based on

  6. SELF-CALIBRATION OF COSMIC MICROWAVE BACKGROUND POLARIZATION EXPERIMENTS

    Energy Technology Data Exchange (ETDEWEB)

    Keating, Brian G.; Yadav, Amit P. S. [Center for Astrophysics and Space Sciences, Department of Physics, University of California, San Diego, 9500 Gilman Drive, La Jolla, CA 92093-0424 (United States); Shimon, Meir [School of Physics and Astronomy, Tel Aviv University, Tel Aviv 69978 (Israel)

    2013-01-10

    Precision measurements of the polarization of the cosmic microwave background (CMB) radiation, especially experiments seeking to detect the odd-parity 'B-modes', have far-reaching implications for cosmology. To detect the B-modes generated during inflation, the flux response and polarization angle of these experiments must be calibrated to exquisite precision. While suitable flux calibration sources abound, polarization angle calibrators are deficient in many respects. Man-made polarized sources are often not located in the antenna's far-field, have spectral properties that are radically different from the CMB's, are cumbersome to implement, and may be inherently unstable over the (long) duration these searches require to detect the faint signature of the inflationary epoch. Astrophysical sources suffer from time, frequency, and spatial variability, are not visible from all CMB observatories, and none are understood with sufficient accuracy to calibrate future CMB polarimeters seeking to probe inflationary energy scales of 10{sup 15} GeV. Both man-made and astrophysical sources require dedicated observations which detract from the amount of integration time usable for detection of the inflationary B-modes. CMB TB and EB modes, expected to identically vanish in the standard cosmological model, can be used to calibrate CMB polarimeters. By enforcing the observed EB and TB power spectra to be consistent with zero, CMB polarimeters can be calibrated to levels not possible with man-made or astrophysical sources. All of this can be accomplished for any polarimeter without any loss of observing time using a calibration source which is spectrally identical to the CMB B-modes.

  7. Characterization of Libyan desert in support of vicarious calibration

    Science.gov (United States)

    Brooke, Bridget L.

    Thorough calibration of electro-optical sensors being designed for today's space-based applications is essential to utilize remotely sensed imagery products. A complete sensor calibration provides in-depth understanding of operation and performance, while verifying that the system meets mission standards. Accurate calibration is crucial in order to give the data physical meaning and allow the imagery to be used and compared for a variety of applications. The use of invariant sites for vicarious calibration has become a valuable tool for the changing design and demands of new instruments. Pseudo-invariant calibration sites provide the opportunity to utilize a much larger site to accommodate the full sensor footprint. Characterizing and understanding these sites is key to successful vicarious calibration. This study focuses on the common pseudo-invariant test site, Libya 4, located in the Saharan desert and two essential atmospheric parameters necessary for characterizing the surface reflectance: column water vapor and aerosol optical depth. These radiometric effects are explored by using a radiative transfer code (e.g., MODTRAN®) to simulate atmospheric changes within realistic environmental ranges and create look up tables to match real world data. We first estimate the water vapor amount in selected data sets using the Continuum Band Interpolated Ratio technique. The aerosol optical depth is then determined using a look up table scheme, with the first estimations of the water vapor amount as inputs to MODTRAN®. Finally, with the atmospheric parameters determined by this iterative approach, we study the surface reflectance of the test site to be utilized in future calibration.

  8. Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Faber, M.H.; Sørensen, John Dalsgaard

    2003-01-01

    . Thereafter the code calibration problem is presented in its principal decision theoretical form and it is discussed how acceptable levels of failure probability (or target reliabilities) may be established. Furthermore suggested values for acceptable annual failure probabilities are given for ultimate...

  9. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Kock, Carsten Weber; Vesth, Allan

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  10. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Vesth, Allan; Kock, Carsten Weber

    The report describes power curve measurements carried out on a given wind turbine. The measurements are carried out in accordance to Ref. [1]. A site calibration has been carried out; see Ref. [2], and the measured flow correction factors for different wind directions are used in the present...

  11. Scalar Calibration of Vector Magnetometers

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz;

    2000-01-01

    the parameters for a given data set. Therefore, a magnetometer may be characterized inexpensively in the Earth's magnetic-field environment. This procedure has been used successfully in the pre-flight calibration of the state-of-the-art magnetometers on board the magnetic mapping satellites Orsted, Astrid-2...

  12. A simple calibration method for mechanically braked cycle ergometers.

    Science.gov (United States)

    Van Praagh, E; Bedu, M; Roddier, P; Coudert, J

    1992-01-01

    The calibration of cycle ergometers should be checked regularly. Some studies have shown calibration errors of more than 40%. A simple, inexpensive calibrating method for mechanically braked cycle ergometers was developed and tried out on a new type of ergocycle. The cycle ergometer was elevated and the crank replaced by a pulley fitted to the shaft. The crank speed (rpm) increased linearly as a function of time when different masses were applied on the pulley. For a given braking force on the cycle ergometer, different accelerations corresponding to the increased pulley forces could be measured. When extrapolating for zero acceleration, it was possible to determine a "limit-force" which allowed the system to be in equilibrium. Additional force creates motion. The same experiments were repeated with increasing braking forces. Using the differently sized gear sprockets of the transmission system, it was possible to calculate the actual force, including all the resistances. The actual force found by the calibrating method was then compared with the indicated force proposed by the manufacturer. With increasing forces, the relative errors decreased from 9.6 to 2.9%. The cycle ergometer calibrated by this technique meets the standards recommended in exercise physiology. PMID:1544728

  13. Automatic calibration of space based manipulators and mechanisms

    Science.gov (United States)

    Everett, Louis J.

    1988-01-01

    Four tasks in manipulator kinematic calibration are summarized. Calibration of a seven degree of freedom manipulator was simulated. A calibration model is presented that can be applied on a closed-loop robot. It is an expansion of open-loop kinematic calibration algorithms subject to constraints. A closed-loop robot with a five-bar linkage transmission was tested. Results show that the algorithm converges within a few iterations. The concept of model differences is formalized. Differences are categorized as structural and numerical, with emphasis on the structural. The work demonstrates that geometric manipulators can be visualized as points in a vector space with the dimension of the space depending solely on the number and type of manipulator joint. Visualizing parameters in a kinematic model as the coordinates locating the manipulator in vector space enables a standard evaluation of the models. Key results include a derivation of the maximum number of parameters necessary for models, a formal discussion on the inclusion of extra parameters, and a method to predetermine a minimum model structure for a kinematic manipulator. A technique is presented that enables single point sensors to gather sufficient information to complete a calibration.

  14. The DICE calibration project: design, characterization, and first results

    CERN Document Server

    Regnault, N; Schahmanèche, K; Guillou, L Le; Antilogus, P; Astier, P; Barrelet, E; Betoule, M; Bongard, S; Cuillandre, J -C; Juramy, C; Pain, R; Rocci, P -F; Tisserand, P; Villa, F

    2015-01-01

    We describe the design, operation, and first results of a photometric calibration project, called DICE (Direct Illumination Calibration Experiment), aiming at achieving precise instrumental calibration of optical telescopes. The heart of DICE is an illumination device composed of 24 narrow-spectrum, high-intensity, light-emitting diodes (LED) chosen to cover the ultraviolet-to-near-infrared spectral range. It implements a point-like source placed at a finite distance from the telescope entrance pupil, yielding a flat field illumination that covers the entire field of view of the imager. The purpose of this system is to perform a lightweight routine monitoring of the imager passbands with a precision better than 5 per-mil on the relative passband normalisations and about 3{\\AA} on the filter cutoff positions. The light source is calibrated on a spectrophotometric bench. As our fundamental metrology standard, we use a photodiode calibrated at NIST. The radiant intensity of each beam is mapped, and spectra are m...

  15. Design, manufacture, and calibration of infrared radiometric blackbody sources

    Energy Technology Data Exchange (ETDEWEB)

    Byrd, D.A.; Michaud, F.D.; Bender, S.C. [and others

    1996-04-01

    A Radiometric Calibration Station (RCS) is being assembled at the Los Alamos National Laboratories (LANL) which will allow for calibration of sensors with detector arrays having spectral capability from about 0.4-15 {mu}m. The configuration of the LANL RCS. Two blackbody sources have been designed to cover the spectral range from about 3-15 {mu}m, operating at temperatures ranging from about 180-350 K within a vacuum environment. The sources are designed to present a uniform spectral radiance over a large area to the sensor unit under test. The thermal uniformity requirement of the blackbody cavities has been one of the key factors of the design, requiring less than 50 mK variation over the entire blackbody surface to attain effective emissivity values of about 0.999. Once the two units are built and verified to the level of about 100 mK at LANL, they will be sent to the National Institute of Standards and Technology (NIST), where at least a factor of two improvement will be calibrated into the blackbody control system. The physical size of these assemblies will require modifications of the existing NIST Low Background Infrared (LBIR) Facility. LANL has constructed a bolt-on addition to the LBIR facility that will allow calibration of our large aperture sources. Methodology for attaining the two blackbody sources at calibrated levels of performance equivalent to present state of the art will be explained in the following.

  16. Precision Spectrophotometric Calibration System for Dark Energy Instruments

    Energy Technology Data Exchange (ETDEWEB)

    Schubnell, Michael S.

    2015-06-30

    For this research we build a precision calibration system and carried out measurements to demonstrate the precision that can be achieved with a high precision spectrometric calibration system. It was shown that the system is capable of providing a complete spectrophotometric calibration at the sub-pixel level. The calibration system uses a fast, high precision monochromator that can quickly and efficiently scan over an instrument’s entire spectral range with a spectral line width of less than 0.01 nm corresponding to a fraction of a pixel on the CCD. The system was extensively evaluated in the laboratory. Our research showed that a complete spectrophotometric calibration standard for spectroscopic survey instruments such as DESI is possible. The monochromator precision and repeatability to a small fraction of the DESI spectrograph LSF was demonstrated with re-initialization on every scan and thermal drift compensation by locking to multiple external line sources. A projector system that mimics telescope aperture for point source at infinity was demonstrated.

  17. Radio Interferometric Calibration Using a Riemannian Manifold

    CERN Document Server

    Yatawatta, Sarod

    2013-01-01

    In order to cope with the increased data volumes generated by modern radio interferometers such as LOFAR (Low Frequency Array) or SKA (Square Kilometre Array), fast and efficient calibration algorithms are essential. Traditional radio interferometric calibration is performed using nonlinear optimization techniques such as the Levenberg-Marquardt algorithm in Euclidean space. In this paper, we reformulate radio interferometric calibration as a nonlinear optimization problem on a Riemannian manifold. The reformulated calibration problem is solved using the Riemannian trust-region method. We show that calibration on a Riemannian manifold has faster convergence with reduced computational cost compared to conventional calibration in Euclidean space.

  18. SDSS-IV/MaNGA: SPECTROPHOTOMETRIC CALIBRATION TECHNIQUE

    Energy Technology Data Exchange (ETDEWEB)

    Yan, Renbin; Sánchez-Gallego, José R. [Department of Physics and Astronomy, University of Kentucky, 505 Rose St., Lexington, KY 40506-0057 (United States); Tremonti, Christy; Bershady, Matthew A.; Eigenbrot, Arthur; Wake, David A. [Department of Astronomy, University of Winsconsin-Madison, 475 N. Charter Street, Madison, WI 53706-1582 (United States); Law, David R. [Space Telescope Science Institute, 3700 San Martin Drive, Baltimore, MD 21218 (United States); Schlegel, David J. [Physics Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720-8160 (United States); Bundy, Kevin [Kavli IPMU (WPI), UTIAS, The University of Tokyo, Kashiwa, Chiba 277-8583 (Japan); Drory, Niv [McDonald Observatory, Department of Astronomy, University of Texas at Austin, 1 University Station, Austin, TX 78712-0259 (United States); MacDonald, Nicholas [Department of Astronomy, Box 351580, University of Washington, Seattle, WA 98195 (United States); Bizyaev, Dmitry [Apache Point Observatory, P.O. Box 59, sunspot, NM 88349 (United States); Blanc, Guillermo A. [Departamento de Astronomía, Universidad de Chile, Camino el Observatorio 1515, Las Condes, Santiago (Chile); Blanton, Michael R.; Hogg, David W. [Center for Cosmology and Particle Physics, Department of Physics, New York University, 4 Washington Place, New York, NY 10003 (United States); Cherinka, Brian [Dunlap Institute for Astronomy and Astrophysics, University of Toronto, 50 St. George Street, Toronto, Ontario M5S 3H4 (Canada); Gunn, James E. [Department of Astrophysical Sciences, Princeton University, Princeton, NJ 08544 (United States); Harding, Paul [Department of Astronomy, Case Western Reserve University, Cleveland, OH 44106 (United States); Sánchez, Sebastian F., E-mail: yanrenbin@uky.edu [Instituto de Astronomia, Universidad Nacional Autonoma de Mexico, A.P. 70-264, 04510 Mexico D.F. (Mexico); and others

    2016-01-15

    Mapping Nearby Galaxies at Apache Point Observatory (MaNGA), one of three core programs in the Sloan Digital Sky Survey-IV, is an integral-field spectroscopic survey of roughly 10,000 nearby galaxies. It employs dithered observations using 17 hexagonal bundles of 2″ fibers to obtain resolved spectroscopy over a wide wavelength range of 3600–10300 Å. To map the internal variations within each galaxy, we need to perform accurate spectral surface photometry, which is to calibrate the specific intensity at every spatial location sampled by each individual aperture element of the integral field unit. The calibration must correct only for the flux loss due to atmospheric throughput and the instrument response, but not for losses due to the finite geometry of the fiber aperture. This requires the use of standard star measurements to strictly separate these two flux loss factors (throughput versus geometry), a difficult challenge with standard single-fiber spectroscopy techniques due to various practical limitations. Therefore, we developed a technique for spectral surface photometry using multiple small fiber-bundles targeting standard stars simultaneously with galaxy observations. We discuss the principles of our approach and how they compare to previous efforts, and we demonstrate the precision and accuracy achieved. MaNGA's relative calibration between the wavelengths of Hα and Hβ has an rms of 1.7%, while that between [N ii] λ6583 and [O ii] λ3727 has an rms of 4.7%. Using extinction-corrected star formation rates and gas-phase metallicities as an illustration, this level of precision guarantees that flux calibration errors will be sub-dominant when estimating these quantities. The absolute calibration is better than 5% for more than 89% of MaNGA's wavelength range.

  19. SDSS-IV/MaNGA: Spectrophotometric Calibration Technique

    Science.gov (United States)

    Yan, Renbin; Tremonti, Christy; Bershady, Matthew A.; Law, David R.; Schlegel, David J.; Bundy, Kevin; Drory, Niv; MacDonald, Nicholas; Bizyaev, Dmitry; Blanc, Guillermo A.; Blanton, Michael R.; Cherinka, Brian; Eigenbrot, Arthur; Gunn, James E.; Harding, Paul; Hogg, David W.; Sánchez-Gallego, José R.; Sánchez, Sebastian F.; Wake, David A.; Weijmans, Anne-Marie; Xiao, Ting; Zhang, Kai

    2016-01-01

    Mapping Nearby Galaxies at Apache Point Observatory (MaNGA), one of three core programs in the Sloan Digital Sky Survey-IV, is an integral-field spectroscopic survey of roughly 10,000 nearby galaxies. It employs dithered observations using 17 hexagonal bundles of 2″ fibers to obtain resolved spectroscopy over a wide wavelength range of 3600-10300 Å. To map the internal variations within each galaxy, we need to perform accurate spectral surface photometry, which is to calibrate the specific intensity at every spatial location sampled by each individual aperture element of the integral field unit. The calibration must correct only for the flux loss due to atmospheric throughput and the instrument response, but not for losses due to the finite geometry of the fiber aperture. This requires the use of standard star measurements to strictly separate these two flux loss factors (throughput versus geometry), a difficult challenge with standard single-fiber spectroscopy techniques due to various practical limitations. Therefore, we developed a technique for spectral surface photometry using multiple small fiber-bundles targeting standard stars simultaneously with galaxy observations. We discuss the principles of our approach and how they compare to previous efforts, and we demonstrate the precision and accuracy achieved. MaNGA's relative calibration between the wavelengths of Hα and Hβ has an rms of 1.7%, while that between [N ii] λ6583 and [O ii] λ3727 has an rms of 4.7%. Using extinction-corrected star formation rates and gas-phase metallicities as an illustration, this level of precision guarantees that flux calibration errors will be sub-dominant when estimating these quantities. The absolute calibration is better than 5% for more than 89% of MaNGA's wavelength range.

  20. Primary Standards Laboratory report, 2nd half 1993

    Energy Technology Data Exchange (ETDEWEB)

    1994-05-01

    The Primary Standards Laboratory (PSL) operates a system-wide primary standards and calibration program for the US Department of Energy, Albuquerque Field Office (DOE/AL). The PSL mission is as follows: to develop and maintain primary standards; to calibrate electrical, physical, and radiation reference standards for customer laboratories (DOE/AL nuclear weapon contractors); to conduct the technical surveys and measurement audits of these laboratories; and to recommend and implement system-wide improvements. This report summarizes activities of the PSL for the second half of 1993 and provides information pertinent to the operation of the DOE/AL Standards and Calibration Program. Specific areas covered include development projects, improvement projects, calibration and special measurements, surveys and audits, customer service, and significant events. Appendixes include certifications and reports;; a discussion about commercial calibration laboratories; PSL memoranda (PSLM); test numbers from the National Institute of Standards and Technology (NIST), formerly the National Bureau of Standards (NBS); and DOE/PSL memoranda on the Standards and Calibration Program with emphasis on traceability of PSL calibrations.

  1. Radioactive contamination in monitors received for calibration

    International Nuclear Information System (INIS)

    The Calibration Laboratory - LABCAL, from the Research Center for Metrology and Testing - METROBRAS, MRA Comercio de Instrumentos Eletronicos Ltda., began activities in October 2008 and, in August 2009, decided to establish a procedure for monitoring tests, external and internal, of all packages received from customers, containing instruments for calibration. The aim was to investigate possible contamination radioactive on these instruments. On July 2011, this procedure was extended to packagings of personal thermoluminescent dosemeters - TLD, received by the newly created Laboratory Laboratorio de Dosimetria Pessoal - LDP . In the monitoring procedure were used monitors with external probe, type pancake, MRA brand, models GP - 500 and MIR 7028. During the 37 months in which this investigation was conducted, were detected 42 cases of radioactive contamination, with the following characteristics: 1) just one case was personal dosimeter, TLD type; 2) just one case was not from a packing from nuclear medicine service - was from a mining company; 3) contamination occurred on packs and instruments, located and/or widespread; 4) contamination values ranged from slightly above the level of background radiation to about a thousand fold. Although METROBRAS has facilities for decontamination, in most cases, especially those of higher contamination, the procedure followed was to store the contaminated material in a room used for storage of radioactive sources. Periodically, each package and/or instrument was monitored, being released when the radiation level matched the background radiation. Every contamination detected, the client and/or owner of the instrument was informed. The Brazilian National Energy Commission - CNEN, was informed, during your public consultation for reviewing the standard for nuclear medicine services, held in mid-2012, having received from METROBRAS the statistical data available at the time. The high frequency of contamination detected and the high

  2. Calibration setting numbers for dose calibrators for the PET isotopes (52)Mn, (64)Cu, (76)Br, (86)Y, (89)Zr, (124)I.

    Science.gov (United States)

    Wooten, A Lake; Lewis, Benjamin C; Szatkowski, Daniel J; Sultan, Deborah H; Abdin, Kinda I; Voller, Thomas F; Liu, Yongjian; Lapi, Suzanne E

    2016-07-01

    For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels.

  3. Radiometric Calibration of Osmi Imagery Using Solar Calibration

    Science.gov (United States)

    Lee, Dong-Han; Kim, Yong-Seung

    2000-12-01

    OSMI (Ocean Scanning Multi-Spectral Imager) raw image data (Level 0) were acquired and radiometrically corrected. We have applied two methods, using solar & dark calibration data from OSMI sensor and comparing with the SeaWiFS data, to the radiometric correction of OSMI raw image data. First, we could get the values of the gain and the offset for each pixel and each band from comparing the solar & dark calibration data with the solar input radiance values, calculated from the transmittance, BRDF (Bidirectional Reflectance Distribution Function) and the solar incidence angle (¥â,¥è) of OSMI sensor. Applying this calibration data to OSMI raw image data, we got the two odd results, the lower value of the radiometric corrected image data than the expected value, and the Venetian Blind Effect in the radiometric corrected image data. Second, we could get the reasonable results from comparing OSMI raw image data with the SeaWiFS data, and get a new problem of OSMI sensor.

  4. SMAP RADAR Calibration and Validation

    Science.gov (United States)

    West, R. D.; Jaruwatanadilok, S.; Chaubel, M. J.; Spencer, M.; Chan, S. F.; Chen, C. W.; Fore, A.

    2015-12-01

    The Soil Moisture Active Passive (SMAP) mission launched on Jan 31, 2015. The mission employs L-band radar and radiometer measurements to estimate soil moisture with 4% volumetric accuracy at a resolution of 10 km, and freeze-thaw state at a resolution of 1-3 km. Immediately following launch, there was a three month instrument checkout period, followed by six months of level 1 (L1) calibration and validation. In this presentation, we will discuss the calibration and validation activities and results for the L1 radar data. Early SMAP radar data were used to check commanded timing parameters, and to work out issues in the low- and high-resolution radar processors. From April 3-13 the radar collected receive only mode data to conduct a survey of RFI sources. Analysis of the RFI environment led to a preferred operating frequency. The RFI survey data were also used to validate noise subtraction and scaling operations in the radar processors. Normal radar operations resumed on April 13. All radar data were examined closely for image quality and calibration issues which led to improvements in the radar data products for the beta release at the end of July. Radar data were used to determine and correct for small biases in the reported spacecraft attitude. Geo-location was validated against coastline positions and the known positions of corner reflectors. Residual errors at the time of the beta release are about 350 m. Intra-swath biases in the high-resolution backscatter images are reduced to less than 0.3 dB for all polarizations. Radiometric cross-calibration with Aquarius was performed using areas of the Amazon rain forest. Cross-calibration was also examined using ocean data from the low-resolution processor and comparing with the Aquarius wind model function. Using all a-priori calibration constants provided good results with co-polarized measurements matching to better than 1 dB, and cross-polarized measurements matching to about 1 dB in the beta release. During the

  5. QA experience at the University of Wisconsin accredited dosimetry calibration laboratory

    Energy Technology Data Exchange (ETDEWEB)

    DeWard, L.A.; Micka, J.A. [Univ. of Wisconsin, Madison, WI (United States)

    1993-12-31

    The University of Wisconsin Accredited Dosimetry Calibration Laboratory (UW ADCL) employs procedure manuals as part of its Quality Assurance (QA) program. One of these manuals covers the QA procedures and results for all of the UW ADCL measurement equipment. The QA procedures are divided into two main areas: QA for laboratory equipment and QA for external chambers sent for calibration. All internal laboratory equipment is checked and recalibrated on an annual basis, after establishing its consistency on a 6-month basis. QA for external instruments involves checking past calibration history as well as comparing to a range of calibration values for specific instrument models. Generally, the authors find that a chamber will have a variation of less than 0.5 % from previous Co-60 calibration factors, and falls within two standard deviations of previous calibrations. If x-ray calibrations are also performed, the energy response of the chamber is plotted and compared to previous instruments of the same model. These procedures give the authors confidence in the transfer of calibration values from National Institute of Standards and Technology (NIST).

  6. Improving integrity of on-line grammage measurement with traceable basic calibration.

    Science.gov (United States)

    Kangasrääsiö, Juha

    2010-07-01

    The automatic control of grammage (basis weight) in paper and board production is based upon on-line grammage measurement. Furthermore, the automatic control of other quality variables such as moisture, ash content and coat weight, may rely on the grammage measurement. The integrity of Kr-85 based on-line grammage measurement systems was studied, by performing basic calibrations with traceably calibrated plastic reference standards. The calibrations were performed according to the EN ISO/IEC 17025 standard, which is a requirement for calibration laboratories. The observed relative measurement errors were 3.3% in the first time calibrations at the 95% confidence level. With the traceable basic calibration method, however, these errors can be reduced to under 0.5%, thus improving the integrity of on-line grammage measurements. Also a standardised algorithm, based on the experience from the performed calibrations, is proposed to ease the adjustment of the different grammage measurement systems. The calibration technique can basically be applied to all beta-radiation based grammage measurements.

  7. [In-Flight Radiometric Calibration for ZY-3 Satellite Multispectral Sensor by Modified Reflectance-Based Method].

    Science.gov (United States)

    Han, Jie; Xie, Yong; Gu, Xing-fa; Yu, Tao; Liu, Qi-yue; Gao, Rong-jun

    2015-03-01

    Through integrating multi-spectral sensor characteristics of ZY-3 satellite, a modified reflectance-based method is proposed and used to achieve ZY-3 satellite multispectral sensor in-flight radiometric calibration. This method chooses level 1A image as data source and establishes geometric model to get an accurate observation geometric parameters at calibration site according to the information provided in image auxiliary documentation, which can reduce the influences on the calibration accuracy from image resampling and observation geometry errors. We use two-point and multi-points methods to calculate the absolute radiometric calibration coefficients of ZY-3 satellite multispectral sensor based on the large campaign at Dongying city, Shan Dong province. Compared with ZY-3 official calibration coefficients, multi-points method has higher accuracy than two-point method. Through analyzing the dispersion between each calibration point and the fitting line, we find that the residual error of water calibration site is larger than others, which of green band is approximately 67.39%. Treating water calibration site as an error, we filter it out using 95.4% confidence level as standard and recalculate the calibration coefficients with multi-points method. The final calibration coefficients show that the relative differences of the first three bands are less than 2% and the last band is less than 5%, which manifests that the proposed radiometric calibration method can obtain accurate and reliable calibration coefficients and is useful for other similar satellites in future.

  8. Calibration transfer via an extreme learning machine auto-encoder.

    Science.gov (United States)

    Chen, Wo-Ruo; Bin, Jun; Lu, Hong-Mei; Zhang, Zhi-Min; Liang, Yi-Zeng

    2016-03-21

    In order to solve the spectra standardization problem in near-infrared (NIR) spectroscopy, a Transfer via Extreme learning machine Auto-encoder Method (TEAM) has been proposed in this study. A comparative study among TEAM, piecewise direct standardization (PDS), generalized least squares (GLS) and calibration transfer methods based on canonical correlation analysis (CCA) was conducted, and the performances of these algorithms were benchmarked with three spectral datasets: corn, tobacco and pharmaceutical tablet spectra. The results show that TEAM is a stable method and can significantly reduce prediction errors compared with PDS, GLS and CCA. TEAM can also achieve the best RMSEPs in most cases with a small number of calibration sets. TEAM is implemented in Python language and available as an open source package at https://github.com/zmzhang/TEAM. PMID:26846329

  9. Attitude Sensor and Gyro Calibration for Messenger

    Science.gov (United States)

    O'Shaughnessy, Daniel; Pittelkau, Mark E.

    2007-01-01

    The Redundant Inertial Measurement Unit Attitude Determination/Calibration (RADICAL(TM)) filter was used to estimate star tracker and gyro calibration parameters using MESSENGER telemetry data from three calibration events. We present an overview of the MESSENGER attitude sensors and their configuration is given, the calibration maneuvers are described, the results are compared with previous calibrations, and variations and trends in the estimated calibration parameters are examined. The warm restart and covariance bump features of the RADICAL(TM) filter were used to estimate calibration parameters from two disjoint telemetry streams. Results show that the calibration parameters converge faster with much less transient variation during convergence than when the filter is cold-started at the start of each telemetry stream.

  10. Crop physiology calibration in the CLM

    Directory of Open Access Journals (Sweden)

    I. Bilionis

    2015-04-01

    scalable and adaptive scheme based on sequential Monte Carlo (SMC. The model showed significant improvement of crop productivity with the new calibrated parameters. We demonstrate that the calibrated parameters are applicable across alternative years and different sites.

  11. Calibration and Validation of Measurement System

    DEFF Research Database (Denmark)

    Kofoed, Jens Peter; Riemann, Sven; Knapp, Wilfried

    The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype.......The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype....

  12. Astrid-2 EMMA Magnetic Calibration

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Risbo, Torben;

    1998-01-01

    The Swedish micro-satellite Astrid-2 contains a tri-axial fluxgate magnetometer with the sensor co-located with a Technical University of Denmark (DTU) star camera for absolute attitude, and extended about 0.9 m on a hinged boom. The magnetometer is part of the RIT EMMA electric and magnetic fields...... experiment built as a collaboration between the DTU, Department of Automation and the Department of Plasma Physics, The Alfvenlaboratory, Royal Institute of Technology (RIT), Stockholm. The final magnetic calibration of the Astrid-2 satellite was done at the Lovoe Magnetic Observatory under the Geological...... Survey of Sweden near Stockholm on the night of May 15.-16., 1997. The magnetic calibration and the intercalibration between the star camera and the magnetic sensor was performed by measuring the Earth's magnetic field and simultaneously observing the star sky with the camera. The rotation matrix between...

  13. Calibration of a Parallel Kinematic Machine Tool

    Institute of Scientific and Technical Information of China (English)

    HE Xiao-mei; DING Hong-sheng; FU Tie; XIE Dian-huang; XU Jin-zhong; LI Hua-feng; LIU Hui-lin

    2006-01-01

    A calibration method is presented to enhance the static accuracy of a parallel kinematic machine tool by using a coordinate measuring machine and a laser tracker. According to the established calibration model and the calibration experiment, the factual 42 kinematic parameters of BKX-I parallel kinematic machine tool are obtained. By circular tests the comparison is made between the calibrated and the uncalibrated parameters and shows that there is 80% improvement in accuracy of this machine tool.

  14. Binary Classifier Calibration: Non-parametric approach

    OpenAIRE

    Naeini, Mahdi Pakdaman; Cooper, Gregory F.; Hauskrecht, Milos

    2014-01-01

    Accurate calibration of probabilistic predictive models learned is critical for many practical prediction and decision-making tasks. There are two main categories of methods for building calibrated classifiers. One approach is to develop methods for learning probabilistic models that are well-calibrated, ab initio. The other approach is to use some post-processing methods for transforming the output of a classifier to be well calibrated, as for example histogram binning, Platt scaling, and is...

  15. GREAT/SOFIA atmospheric calibration

    OpenAIRE

    Guan, Xin; Stutzki, Jürgen; Graf, Urs U.; Güsten, Rolf; Okada, Yoko; Torres, Miguel Angel Requena; Simon, Robert; Wiesemeyer, Helmut

    2012-01-01

    The GREAT observations need frequency-selective calibration across the passband for the residual atmospheric opacity at flight altitude. At these altitudes the atmospheric opacity has both narrow and broad spectral features. To determine the atmospheric transmission at high spectral resolution, GREAT compares the observed atmospheric emission with atmospheric model predictions, and therefore depends on the validity of the atmospheric models. We discusse the problems identified in this compari...

  16. Calibrating thermal behavior of electronics

    Science.gov (United States)

    Chainer, Timothy J.; Parida, Pritish R.; Schultz, Mark D.

    2016-05-31

    A method includes determining a relationship between indirect thermal data for a processor and a measured temperature associated with the processor, during a calibration process, obtaining the indirect thermal data for the processor during actual operation of the processor, and determining an actual significant temperature associated with the processor during the actual operation using the indirect thermal data for the processor during actual operation of the processor and the relationship.

  17. Calibration of Robot Reference Frames for Enhanced Robot Positioning Accuracy

    OpenAIRE

    Cheng, Frank Shaopeng

    2008-01-01

    This chapter discussed the importance and methods of conducting robot workcell calibration for enhancing the accuracy of the robot TCP positions in industrial robot applications. It shows that the robot frame transformations define the robot geometric parameters such as joint position variables, link dimensions, and joint offsets in an industrial robot system. The D-H representation allows the robot designer to model the robot motion geometry with the four standard D-H parameters. The robot k...

  18. Embodying, calibrating and caring for a local model of obesity

    DEFF Research Database (Denmark)

    Winther, Jonas; Hillersdal, Line

    Interdisciplinary research collaborations are increasingly made a mandatory 'standard' within strategic research grants. Collaborations between the natural, social and humanistic sciences are conceptualized as uniquely suited to study pressing societal problems. The obesity epidemic has been...... is an ongoing process of configuring but also extending beyond already established models of obesity. We argue that an articulation of such practices of local care, embodiment and calibration are crucial for the appreciation, evaluation and transferability of interdisciplinary obesity research....

  19. Unassisted 3D camera calibration

    Science.gov (United States)

    Atanassov, Kalin; Ramachandra, Vikas; Nash, James; Goma, Sergio R.

    2012-03-01

    With the rapid growth of 3D technology, 3D image capture has become a critical part of the 3D feature set on mobile phones. 3D image quality is affected by the scene geometry as well as on-the-device processing. An automatic 3D system usually assumes known camera poses accomplished by factory calibration using a special chart. In real life settings, pose parameters estimated by factory calibration can be negatively impacted by movements of the lens barrel due to shaking, focusing, or camera drop. If any of these factors displaces the optical axes of either or both cameras, vertical disparity might exceed the maximum tolerable margin and the 3D user may experience eye strain or headaches. To make 3D capture more practical, one needs to consider unassisted (on arbitrary scenes) calibration. In this paper, we propose an algorithm that relies on detection and matching of keypoints between left and right images. Frames containing erroneous matches, along with frames with insufficiently rich keypoint constellations, are detected and discarded. Roll, pitch yaw , and scale differences between left and right frames are then estimated. The algorithm performance is evaluated in terms of the remaining vertical disparity as compared to the maximum tolerable vertical disparity.

  20. Model Calibration in Watershed Hydrology

    Science.gov (United States)

    Yilmaz, Koray K.; Vrugt, Jasper A.; Gupta, Hoshin V.; Sorooshian, Soroosh

    2009-01-01

    Hydrologic models use relatively simple mathematical equations to conceptualize and aggregate the complex, spatially distributed, and highly interrelated water, energy, and vegetation processes in a watershed. A consequence of process aggregation is that the model parameters often do not represent directly measurable entities and must, therefore, be estimated using measurements of the system inputs and outputs. During this process, known as model calibration, the parameters are adjusted so that the behavior of the model approximates, as closely and consistently as possible, the observed response of the hydrologic system over some historical period of time. This Chapter reviews the current state-of-the-art of model calibration in watershed hydrology with special emphasis on our own contributions in the last few decades. We discuss the historical background that has led to current perspectives, and review different approaches for manual and automatic single- and multi-objective parameter estimation. In particular, we highlight the recent developments in the calibration of distributed hydrologic models using parameter dimensionality reduction sampling, parameter regularization and parallel computing.

  1. ['Gold standard', not 'golden standard'

    NARCIS (Netherlands)

    Claassen, J.A.H.R.

    2005-01-01

    In medical literature, both 'gold standard' and 'golden standard' are employed to describe a reference test used for comparison with a novel method. The term 'gold standard' in its current sense in medical research was coined by Rudd in 1979, in reference to the monetary gold standard. In the same w

  2. New method for calibration of sun photometers

    Institute of Scientific and Technical Information of China (English)

    H.H.Asadov; I.G.Chobanzadeh

    2009-01-01

    A new method for calibration of sun photometers based on Bouguer-Beer law is proposed.The developed basic equation of calibration makes it possible to formulate the derivative methods of calibration on the basis of photometric measurements upon optical air masses,the ratio of which is an integer number.

  3. 48 CFR 908.7113 - Calibration services.

    Science.gov (United States)

    2010-10-01

    ... 48 Federal Acquisition Regulations System 5 2010-10-01 2010-10-01 false Calibration services. 908... ACQUISITION PLANNING REQUIRED SOURCES OF SUPPLIES AND SERVICES Acquisition of Special Items 908.7113 Calibration services. Orders for calibration services may be placed with the National Institute of...

  4. Net analyte signal calculation for multivariate calibration

    NARCIS (Netherlands)

    Ferre, J.; Faber, N.M.

    2003-01-01

    A unifying framework for calibration and prediction in multivariate calibration is shown based on the concept of the net analyte signal (NAS). From this perspective, the calibration step can be regarded as the calculation of a net sensitivity vector, whose length is the amount of net signal when the

  5. Design of an ultra-portable field transfer radiometer supporting automated vicarious calibration

    Science.gov (United States)

    Anderson, Nikolaus; Thome, Kurtis; Czapla-Myers, Jeffrey; Biggar, Stuart

    2015-09-01

    The University of Arizona Remote Sensing Group (RSG) began outfitting the radiometric calibration test site (RadCaTS) at Railroad Valley Nevada in 2004 for automated vicarious calibration of Earth-observing sensors. RadCaTS was upgraded to use RSG custom 8-band ground viewing radiometers (GVRs) beginning in 2011 and currently four GVRs are deployed providing an average reflectance for the test site. This measurement of ground reflectance is the most critical component of vicarious calibration using the reflectance-based method. In order to ensure the quality of these measurements, RSG has been exploring more efficient and accurate methods of on-site calibration evaluation. This work describes the design of, and initial results from, a small portable transfer radiometer for the purpose of GVR calibration validation on site. Prior to deployment, RSG uses high accuracy laboratory calibration methods in order to provide radiance calibrations with low uncertainties for each GVR. After deployment, a solar radiation based calibration has typically been used. The method is highly dependent on a clear, stable atmosphere, requires at least two people to perform, is time consuming in post processing, and is dependent on several large pieces of equipment. In order to provide more regular and more accurate calibration monitoring, the small portable transfer radiometer is designed for quick, one-person operation and on-site field calibration comparison results. The radiometer is also suited for laboratory calibration use and thus could be used as a transfer radiometer calibration standard for ground viewing radiometers of a RadCalNet site.

  6. Calibration of visual model for space manipulator with a hybrid LM-GA algorithm

    Science.gov (United States)

    Jiang, Wensong; Wang, Zhongyu

    2016-01-01

    A hybrid LM-GA algorithm is proposed to calibrate the camera system of space manipulator to improve its locational accuracy. This algorithm can dynamically fuse the Levenberg-Marqurdt (LM) algorithm and Genetic Algorithm (GA) together to minimize the error of nonlinear camera model. LM algorithm is called to optimize the initial camera parameters that are generated by genetic process previously. Iteration should be stopped if the optimized camera parameters meet the accuracy requirements. Otherwise, new populations are generated again by GA and optimized afresh by LM algorithm until the optimal solutions meet the accuracy requirements. A novel measuring machine of space manipulator is designed to on-orbit dynamic simulation and precision test. The camera system of space manipulator, calibrated by hybrid LM-GA algorithm, is used for locational precision test in this measuring instrument. The experimental results show that the mean composite errors are 0.074 mm for hybrid LM-GA camera calibration model, 1.098 mm for LM camera calibration model, and 1.202 mm for GA camera calibration model. Furthermore, the composite standard deviations are 0.103 mm for the hybrid LM-GA camera calibration model, 1.227 mm for LM camera calibration model, and 1.351 mm for GA camera calibration model. The accuracy of hybrid LM-GA camera calibration model is more than 10 times higher than that of other two methods. All in all, the hybrid LM-GA camera calibration model is superior to both the LM camera calibration model and GA camera calibration model.

  7. DIGITAL CAMERA CALIBRATION USING IMAGES TAKEN FROM AN UNMANNED AERIAL VEHICLE

    Directory of Open Access Journals (Sweden)

    M. Pérez

    2012-09-01

    Full Text Available For calibrating the camera, an accurate determination of the interior orientation parameters is needed. For more accurate results, the calibration images should be taken under conditions that are similar to the field samples. The aim of this work is the establishment of an efficient and accurate digital camera calibration method to be used in particular working conditions, as it can be found with our UAV (Unmanned Aerial Vehicle photogrammetric projects. The UAV used in this work was md4-200 modelled by Microdrones. The microdrone is also equipped with a standard digital non- metric camera, the Pentax Optio A40 camera. To find out the interior orientation parameters of the digital camera, two calibration methods were done. A lab calibration based on a flat pattern and a field calibration were fulfilled. To carry out the calibration, Photomodeler Scanner software was used in both cases. The lab calibration process was completely automatic using a calibration grid. The focal length was fixed at widest angle and the network included a total of twelve images with± 90º roll angles. In order to develop the field calibration, a flight plan was programmed including a total of twelve images. In the same way as in the lab calibration, the focal length was fixed at widest angle. The field test used in the study was a flat surface located on the University of Almería campus and a set of 67 target points were placed. The calibration field area was 25 × 25 m approximately and the altitude flight over ground was 50 m. After the software processing, the camera calibration parameter values were obtained. The paper presents the process, the results and the accuracy of these calibration methods. The field calibration method reduced the final total error obtained in the previous lab calibration. Furthermore the overall RMSs obtained from both methods are similar. Therefore we will apply the field calibration results to all our photogrammetric projects in which

  8. Perfecting the Photometric Calibration of the ACS CCD Cameras

    Science.gov (United States)

    Bohlin, Ralph C.

    2016-09-01

    Newly acquired data and improved data reduction algorithms mandate a fresh look at the absolute flux calibration of the charge-coupled device cameras on the Hubble Space Telescope (HST) Advanced Camera for Surveys (ACS). The goals are to achieve a 1% accuracy and to make this calibration more accessible to the HST guest investigator. Absolute fluxes from the CALSPEC1 database for three primary hot 30,000–60,000K WDs define the sensitivity calibrations for the Wide Field Channel (WFC) and High Resolution Channel (HRC) filters. The external uncertainty for the absolute flux is ˜1%, while the internal consistency of the sensitivities in the broadband ACS filters is ˜0.3% among the three primary WD flux standards. For stars as cool as K type, the agreement with the CALSPEC standards is within 1% at the WFC1-1K subarray position, which achieves the 1% precision goal for the first time. After making a small adjustment to the filter bandpass for F814W, the 1% precision goal is achieved over the full F814W WFC field of view for stars of K type and hotter. New encircled energies and absolute sensitivities replace the seminal results of Sirianni et al. that were published in 2005. After implementing the throughput updates, synthetic predictions of the WFC and HRC count rates for the average of the three primary WD standard stars agree with the observations to 0.1%.

  9. Perfecting the Photometric Calibration of the ACS CCD Cameras

    Science.gov (United States)

    Bohlin, Ralph C.

    2016-09-01

    Newly acquired data and improved data reduction algorithms mandate a fresh look at the absolute flux calibration of the charge-coupled device cameras on the Hubble Space Telescope (HST) Advanced Camera for Surveys (ACS). The goals are to achieve a 1% accuracy and to make this calibration more accessible to the HST guest investigator. Absolute fluxes from the CALSPEC1 database for three primary hot 30,000-60,000K WDs define the sensitivity calibrations for the Wide Field Channel (WFC) and High Resolution Channel (HRC) filters. The external uncertainty for the absolute flux is ˜1%, while the internal consistency of the sensitivities in the broadband ACS filters is ˜0.3% among the three primary WD flux standards. For stars as cool as K type, the agreement with the CALSPEC standards is within 1% at the WFC1-1K subarray position, which achieves the 1% precision goal for the first time. After making a small adjustment to the filter bandpass for F814W, the 1% precision goal is achieved over the full F814W WFC field of view for stars of K type and hotter. New encircled energies and absolute sensitivities replace the seminal results of Sirianni et al. that were published in 2005. After implementing the throughput updates, synthetic predictions of the WFC and HRC count rates for the average of the three primary WD standard stars agree with the observations to 0.1%.

  10. Two approaches in scanner-printer calibration: colorimetric space-based vs. “closed-loop&rdquo

    OpenAIRE

    Ostromoukhov, V.; Hersch, R. D.; Peraire, C. (C.); Emmel, P.; Amidror, I.

    1994-01-01

    Studies two different table-based approaches for the calibration of electronic imaging systems. The first approach, which is the classical one, uses the device-independent CIE-XYZ colorimetric space as an intermediate standard space. Input and output devices such as scanners, displays and printers are calibrated separately with respect to the objective CIE-XYZ space. The calibration process requires establishing a 3D mapping between a scanner's device-dependent RGB space and a device-independ...

  11. Scatterometer-Calibrated Stability Verification Method

    Science.gov (United States)

    McWatters, Dalia A.; Cheetham, Craig M.; Huang, Shouhua; Fischman, Mark A.; CHu, Anhua J.; Freedman, Adam P.

    2011-01-01

    The requirement for scatterometer-combined transmit-receive gain variation knowledge is typically addressed by sampling a portion of the transmit signal, attenuating it with a known-stable attenuation, and coupling it into the receiver chain. This way, the gain variations of the transmit and receive chains are represented by this loop-back calibration signal, and can be subtracted from the received remote radar echo. Certain challenges are presented by this process, such as transmit and receive components that are outside of this loop-back path and are not included in this calibration, as well as the impracticality for measuring the transmit and receive chains stability and post fabrication separately, without the resulting measurement errors from the test set up exceeding the requirement for the flight instrument. To cover the RF stability design challenge, the portions of the scatterometer that are not calibrated by the loop-back, (e.g., attenuators, switches, diplexers, couplers, and coaxial cables) are tightly thermally controlled, and have been characterized over temperature to contribute less than 0.05 dB of calibration error over worst-case thermal variation. To address the verification challenge, including the components that are not calibrated by the loop-back, a stable fiber optic delay line (FODL) was used to delay the transmitted pulse, and to route it into the receiver. In this way, the internal loopback signal amplitude variations can be compared to the full transmit/receive external path, while the flight hardware is in the worst-case thermal environment. The practical delay for implementing the FODL is 100 s. The scatterometer pulse width is 1 ms so a test mode was incorporated early in the design phase to scale the 1 ms pulse at 100-Hz pulse repetition interval (PRI), by a factor of 18, to be a 55 s pulse with 556 s PRI. This scaling maintains the duty cycle, thus maintaining a representative thermal state for the RF components. The FODL consists

  12. Preliminary calibration of the ACP safeguards neutron counter

    Science.gov (United States)

    Lee, T. H.; Kim, H. D.; Yoon, J. S.; Lee, S. Y.; Swinhoe, M.; Menlove, H. O.

    2007-10-01

    The Advanced Spent Fuel Conditioning Process (ACP), a kind of pyroprocess, has been developed at the Korea Atomic Energy Research Institute (KAERI). Since there is no IAEA safeguards criteria for this process, KAERI has developed a neutron coincidence counter to make it possible to perform a material control and accounting (MC&A) for its ACP materials for the purpose of a transparency in the peaceful uses of nuclear materials at KAERI. The test results of the ACP Safeguards Neutron Counter (ASNC) show a satisfactory performance for the Doubles count measurement with a low measurement error for its cylindrical sample cavity. The neutron detection efficiency is about 21% with an error of ±1.32% along the axial direction of the cavity. Using two 252Cf neutron sources, we obtained various parameters for the Singles and Doubles rates for the ASNC. The Singles, Doubles, and Triples rates for a 252Cf point source were obtained by using the MCNPX code and the results for the ft8 cap multiplicity tally option with the values of ɛ, fd, and ft measured with a strong source most closely match the measurement results to within a 1% error. A preliminary calibration curve for the ASNC was generated by using the point model equation relationship between 244Cm and 252Cf and the calibration coefficient for the non-multiplying sample is 2.78×10 5 (Doubles counts/s/g 244Cm). The preliminary calibration curves for the ACP samples were also obtained by using an MCNPX simulation. A neutron multiplication influence on an increase of the Doubles rate for a metal ingot and UO2 powder is clearly observed. These calibration curves will be modified and complemented, when hot calibration samples become available. To verify the validity of this calibration curve, a measurement of spent fuel standards for a known 244Cm mass will be performed in the near future.

  13. Foreground-induced biases in CMB polarimeter self-calibration

    Science.gov (United States)

    Abitbol, Maximilian H.; Hill, James; Johnson, Bradley

    2016-06-01

    Precise polarization measurements of the cosmic microwave background (CMB) require accurate knowledge of the instrument orientation relative to the sky frame used to define the cosmological Stokes parameters. Suitable celestial calibration sources that could be used to measure the polarimeter orientation angle are limited, so current experiments commonly `self-calibrate.' The self-calibration method exploits the theoretical fact that the EB and TB cross-spectra of the CMB vanish in the standard cosmological model, so any detected EB and TB signals must be due to systematic errors. However, this assumption neglects the fact that polarized Galactic foregrounds in a given portion of the sky may have non-zero EB and TB cross-spectra. If these foreground signals remain in the observations, then they will bias the self-calibrated telescope polarization angle and produce a spurious B-mode signal. In this paper, we estimate the foreground-induced bias for various instrument configurations and then expand the self-calibration formalism to account for polarized foreground signals. Assuming the EB correlation signal for dust is in the range constrained by angular power spectrum measurements from Planck at 353 GHz (scaled down to 150 GHz), then the bias is negligible for high angular resolution experiments, which have access to CMB-dominated high 'ell' modes with which to self-calibrate. Low-resolution experiments observing particularly dusty sky patches can have a bias as large as 0.5°. A miscalibration of this magnitude generates a spurious BB signal corresponding to a tensor-to-scalar ratio of approximately r ~ 2 × 10-3, within the targeted range of planned experiments.

  14. Improving LSST Photometric Calibration with Gaia Data

    CERN Document Server

    Axelrod, Tim

    2014-01-01

    We consider the possibility that the Gaia mission can supply data which will improve the photometric calibration of LSST. After outlining the LSST calibra- tion process and the information that will be available from Gaia, we explore two options for using Gaia data. The first is to use Gaia G-band photometry of selected stars, in conjunction with knowledge of the stellar parameters Teff, log g, and AV, and in some cases Z, to create photometric standards in the LSST u, g, r, i, z, and y bands. The accuracies of the resulting standard magnitudes are found to be insufficient to satisfy LSST requirements when generated from main sequence (MS) stars, but generally adequate from DA white dwarfs (WD). The second option is combine the LSST bandpasses into a synthetic Gaia G band, which is a close approximation to the real Gaia G band. This allows synthetic Gaia G photometry to be directly compared with actual Gaia G photometry at a level of accuracy which is useful for both verifying and improving LSST photometric c...

  15. Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications

    Directory of Open Access Journals (Sweden)

    Thomas Jerry A

    2008-03-01

    Full Text Available Abstract Background Breast density is a significant breast cancer risk factor. Currently, there is no standard method for measuring this important factor. Work presented here represents an essential component of an ongoing project that seeks to determine the appropriate method for calibrating (standardizing mammography image data to account for the x-ray image acquisition influences. Longer term goals of this project are to make accurate breast density measurements in support of risk studies. Methods Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two full field digital mammography (FFDM systems with breast tissue equivalent phantom imaging and compared. Normalization methods were studied to assess the possibility of reducing the amount of calibration data collection. The percent glandular calibration map functional form was investigated. Spatial variations in the calibration data were used to assess the uncertainty in the calibration application by applying error propagation analyses. Results Logarithmic response curves are well approximated as linear. Measured effective x-ray attenuation coefficients are characteristic quantities independent of the imaging system and are in agreement with those predicted numerically. Calibration data collection can be reduced by applying a simple normalization technique. The calibration map is well approximated as linear. Intrasystem calibration variation was on the order of four percent, which was approximately half of the intersystem variation. Conclusion FFDM systems provide a quantitative output, and the calibration quantities presented here may be used for data acquired on similar FFDM systems.

  16. Calibration of ground-based lidar instrument WLS7-151

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  17. Calibration of ground-based lidar instrument WLS7-213

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  18. Calibration of ground-based Lidar instrument WLS7-226

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  19. Calibration of ground-based Lidar instrument WLS7-34

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  20. Calibration of ground-based Lidar instrument WLS866-5

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  1. Calibration of ground-based Lidar instrument WLS7-66

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  2. Calibration of ground-based Lidar instrument WLS866-8

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  3. Calibration of ground-based Lidar instrument WLS7-218

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  4. Calibration of ground-based lidar instrument WLS7-221

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  5. Calibration of ground-based Lidar instrument WLS7-37

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  6. Calibration of ground-based Lidar instrument WLS7-341

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  7. Calibration of ground-based Lidar instrument WLS7-91

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  8. Calibration of ground-based Lidar instrument WLS7-150

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  9. Calibration of ground-based Lidar instrument WLS7-73

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  10. Calibration of ground-based Lidar instrument WLS7-264

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  11. A likelihood method to cross-calibrate air-shower detectors

    CERN Document Server

    Dembinski, H P; Mariş, I C; Roth, M; Veberič, D

    2016-01-01

    We present a detailed statistical treatment of the energy calibration of hybrid air-shower detectors, which combine a surface detector array and a fluorescence detector, to obtain an unbiased estimate of the calibration curve. The special features of calibration data from air showers prevent unbiased results, if a standard least-squares fit is applied to the problem. We develop a general maximum-likelihood approach, based on the detailed statistical model, to solve the problem. Our approach was developed for the Pierre Auger Observatory, but the applied principles are general and can be transferred to other air-shower experiments, even to the cross-calibration of other observables. Since our general likelihood function is expensive to compute, we derive two approximations with significantly smaller computational cost. In the recent years both have been used to calibrate data of the Pierre Auger Observatory. We demonstrate that these approximations introduce negligible bias when they are applied to simulated t...

  12. Calibration methods of plane-parallel ionization chambers used in electron dosimetry

    International Nuclear Information System (INIS)

    The use of linear accelerators in radiotherapy is of great importance in Medicine, and according to international recommendations the electron beam dosimetry has to be performed using plane-parallel ionization chambers, previously calibrated in standard gamma radiation fields at accredited laboratories. In this work, calibration methods of plane-parallel ionization chambers used in dosimetry procedures of high energy electron beams of clinical accelerators were presented, tested and intercompared. The experiments were carried out using gamma radiation beams of 60 Co at the Calibration Laboratory of Clinical Dosemeters at IPEN and electron beams od 4 to 16 MeV at the Radiotherapy Department of Hospital Israelita Albert Einstein, Sao Paulo. A method was chosen to be established at IPEN. Proposals of the calibration procedure, calibration certificate and data sheets are presented. (author)

  13. Accounting standards

    NARCIS (Netherlands)

    B. Stellinga

    2014-01-01

    The European and global regulation of accounting standards have witnessed remarkable changes over the past twenty years. In the early 1990s, EU accounting practices were fragmented along national lines and US accounting standards were the de facto global standards. Since 2005, all EU listed companie

  14. Muon Energy Calibration of the MINOS Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Miyagawa, Paul S.

    2004-09-01

    MINOS is a long-baseline neutrino oscillation experiment designed to search for conclusive evidence of neutrino oscillations and to measure the oscillation parameters precisely. MINOS comprises two iron tracking calorimeters located at Fermilab and Soudan. The Calibration Detector at CERN is a third MINOS detector used as part of the detector response calibration programme. A correct energy calibration between these detectors is crucial for the accurate measurement of oscillation parameters. This thesis presents a calibration developed to produce a uniform response within a detector using cosmic muons. Reconstruction of tracks in cosmic ray data is discussed. This data is utilized to calculate calibration constants for each readout channel of the Calibration Detector. These constants have an average statistical error of 1.8%. The consistency of the constants is demonstrated both within a single run and between runs separated by a few days. Results are presented from applying the calibration to test beam particles measured by the Calibration Detector. The responses are calibrated to within 1.8% systematic error. The potential impact of the calibration on the measurement of oscillation parameters by MINOS is also investigated. Applying the calibration reduces the errors in the measured parameters by {approx} 10%, which is equivalent to increasing the amount of data by 20%.

  15. "Calibration-on-the-spot": How to calibrate an EMCCD camera from its images.

    Science.gov (United States)

    Mortensen, Kim I; Flyvbjerg, Henrik

    2016-07-06

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information.

  16. Calibration-on-the-spot”: How to calibrate an EMCCD camera from its images

    DEFF Research Database (Denmark)

    Mortensen, Kim; Flyvbjerg, Henrik

    2016-01-01

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction......-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information....

  17. Calibration of the Cherenkov Telescope Array

    CERN Document Server

    Gaug, Markus; Berge, David; Reyes, Raquel de los; Doro, Michele; Foerster, Andreas; Maccarone, Maria Concetta; Parsons, Dan; van Eldik, Christopher

    2015-01-01

    The construction of the Cherenkov Telescope Array is expected to start soon. We will present the baseline methods and their extensions currently foreseen to calibrate the observatory. These are bound to achieve the strong requirements on allowed systematic uncertainties for the reconstructed gamma-ray energy and flux scales, as well as on the pointing resolution, and on the overall duty cycle of the observatory. Onsite calibration activities are designed to include a robust and efficient calibration of the telescope cameras, and various methods and instruments to achieve calibration of the overall optical throughput of each telescope, leading to both inter-telescope calibration and an absolute calibration of the entire observatory. One important aspect of the onsite calibration is a correct understanding of the atmosphere above the telescopes, which constitutes the calorimeter of this detection technique. It is planned to be constantly monitored with state-of-the-art instruments to obtain a full molecular and...

  18. Multi-Axis Accelerometer Calibration System

    Science.gov (United States)

    Finley, Tom; Parker, Peter

    2010-01-01

    A low-cost, portable, and simplified system has been developed that is suitable for in-situ calibration and/or evaluation of multi-axis inertial measurement instruments. This system overcomes facility restrictions and maintains or improves the calibration quality for users of accelerometer-based instruments with applications in avionics, experimental wind tunnel research, and force balance calibration applications. The apparatus quickly and easily positions a multi-axis accelerometer system into a precisely known orientation suitable for in-situ quality checks and calibration. In addition, the system incorporates powerful and sophisticated statistical methods, known as response surface methodology and statistical quality control. These methods improve calibration quality, reduce calibration time, and allow for increased calibration frequency, which enables the monitoring of instrument stability over time.

  19. MODIS Instrument Operation and Calibration Improvements

    Science.gov (United States)

    Xiong, X.; Angal, A.; Madhavan, S.; Link, D.; Geng, X.; Wenny, B.; Wu, A.; Chen, H.; Salomonson, V.

    2014-01-01

    Terra and Aqua MODIS have successfully operated for over 14 and 12 years since their respective launches in 1999 and 2002. The MODIS on-orbit calibration is performed using a set of on-board calibrators, which include a solar diffuser for calibrating the reflective solar bands (RSB) and a blackbody for the thermal emissive bands (TEB). On-orbit changes in the sensor responses as well as key performance parameters are monitored using the measurements of these on-board calibrators. This paper provides an overview of MODIS on-orbit operation and calibration activities, and instrument long-term performance. It presents a brief summary of the calibration enhancements made in the latest MODIS data collection 6 (C6). Future improvements in the MODIS calibration and their potential applications to the S-NPP VIIRS are also discussed.

  20. An Atlas of Spectrophotometric Landolt Standard Stars

    OpenAIRE

    Stritzinger, Maximilian; Suntzeff, Nicholas B.; Hamuy, Mario; Challis, Peter; Demarco, Ricardo; Germany, Lisa; Soderberg, A. M.

    2005-01-01

    We present CCD observations of 102 Landolt standard stars obtained with the R-C spectrograph on the CTIO 1.5 m telescope. Using stellar atmosphere models we have extended the flux points to our six spectrophotometric secondary standards, in both the blue and the red, allowing us to produce flux-calibrated spectra that span a wavelength range from 3050 \\AA to 1.1 \\micron. Mean differences between UBVRI spectrophotometry computed using Bessell's standard passbands and Landolt's published photom...

  1. ANALYSIS OF EPA (ENVIRONMENTAL PROTECTION AGENCY) PROTOCOL GASES USED FOR CALIBRATION AND AUDITS OF CONTINUOUS EMISSION MONITORING SYSTEMS AND AMBIENT AIR ANALYZERS - RESULTS OF AUDIT 6

    Science.gov (United States)

    A performance audit was conducted on EPA Protocol Gases used for calibration and audits of continuous emission monitoring systems and ambient air analyzers. Fifty gaseous pollutant calibraton standards were purchased from eleven specialty gas producers. These standards contained ...

  2. The estimation of total petroleum hydrocarbons content in waste water by IR spectrometry with multivariate calibrations.

    Science.gov (United States)

    Vershinin, Viacheslav I; Petrov, Sergey V

    2016-02-01

    Alkanes, cycloalkanes and arenes have rather different sensitivities to IR-spectrometric determination, leading to high relative uncertainty (δc) for the total petroleum hydrocarbon index (TPH) in natural and waste waters. Another source of TPH uncertainty is the mismatch of group composition of the hydrocarbon mixture in the sample and in the standard substance used for one-dimensional calibration. Increasing the number of wavelengths and using of multivariate calibrations permit the reduction of δc to <10% rel. These calibrations may be constructed from IR-spectra and findings of extracts from aqueous solutions with known content of hydrocarbons. The method takes into account the losses of hydrocarbons during sample preparation. The accuracy of TPH estimations for this method is much better than for standard methods based on one-dimensional calibration with Simard mixture. This new method is useful in produced waste water analysis. PMID:26653437

  3. Calibrations and evaluation of the control program at the National Laboratory during 1995

    International Nuclear Information System (INIS)

    The Dosimetry Laboratory at the Swedish Radiation Protection Institute (SSI) is a National Laboratory for the dosimetric quantities kerma, absorbed dose and dose equivalent. The activity is based on established routines for how calibrations should be made and a control and a calibration program for the used standards. This report gives a brief summary of the calibrations performed during 1995 and a more detailed description and analysis of the control program during this year. To summarize all the controls and calibrations of standards made during the year makes it easier to draw conclusions about the long term stability and possible malfunctions. Therefore, this summary makes an important part of the quality assurance program at the National Laboratory. 10 figs, 24 tabs

  4. Calibrations and evaluation of the control program at the National Laboratory during 1994

    International Nuclear Information System (INIS)

    The Dosimetry Laboratory at the Swedish Radiation Protection Institute (SSI) is a National Laboratory for the dosimetric quantities kerma, absorbed dose and dose equivalent. The activity is based on established routines for how calibrations should be made and a control and calibration program for the used standards. This report gives a brief summary of the calibrations performed during 1994 and a more detailed description and analysis of the control program during this year. To summarize all the controls and calibrations of standards made during the year makes it easier to draw conclusions about the long term stability and possible malfunctions. Therefore, this summary makes an important part of the quality assurance program at the National Laboratory

  5. Precise astronomical flux calibration and its impact on studying the nature of the dark energy

    Science.gov (United States)

    Stubbs, Christopher W.; Brown, Yorke J.

    2015-12-01

    Measurements of the luminosity of Type Ia supernovae versus redshift provided the original evidence for the accelerating expansion of the Universe and the existence of dark energy. Despite substantial improvements in survey methodology, systematic uncertainty in flux calibration dominates the error budget for this technique, exceeding both statistics and other systematic uncertainties. Consequently, any further collection of Type Ia supernova data will fail to refine the constraints on the nature of dark energy unless we also improve the state of the art in astronomical flux calibration to the order of 1%. We describe how these systematic errors arise from calibration of instrumental sensitivity, atmospheric transmission and Galactic extinction, and discuss ongoing efforts to meet the 1% precision challenge using white dwarf stars as celestial standards, exquisitely calibrated detectors as fundamental metrologic standards, and real-time atmospheric monitoring.

  6. Precise Astronomical Flux Calibration and its Impact on Studying the Nature of Dark Energy

    CERN Document Server

    Stubbs, Christopher W

    2016-01-01

    Measurements of the luminosity of type Ia supernovae vs. redshift provided the original evidence for the accelerating expansion of the Universe and the existence of dark energy. Despite substantial improvements in survey methodology, systematic uncertainty in flux calibration dominates the error budget for this technique, exceeding both statistics and other systematic uncertainties. Consequently, any further collection of type Ia supernova data will fail to refine the constraints on the nature of dark energy unless we also improve the state of the art in astronomical flux calibration to the order of 1%. We describe how these systematic errors arise from calibration of instrumental sensitivity, atmospheric transmission, and Galactic extinction, and discuss ongoing efforts to meet the 1% precision challenge using white dwarf stars as celestial standards, exquisitely calibrated detectors as fundamental metrologic standards, and real-time atmospheric monitoring.

  7. ACCURATE KAP METER CALIBRATION AS A PREREQUISITE FOR OPTIMISATION IN PROJECTION RADIOGRAPHY.

    Science.gov (United States)

    Malusek, A; Sandborg, M; Carlsson, G Alm

    2016-06-01

    Modern X-ray units register the air kerma-area product, PKA, with a built-in KAP meter. Some KAP meters show an energy-dependent bias comparable with the maximum uncertainty articulated by the IEC (25 %), adversely affecting dose-optimisation processes. To correct for the bias, a reference KAP meter calibrated at a standards laboratory and two calibration methods described here can be used to achieve an uncertainty of standards laboratory, Q0, to any beam quality, Q, in the clinic. Alternatively, beam quality corrections are measured with an energy-independent dosemeter via a reference beam quality in the clinic, Q1, to beam quality, Q Biases up to 35 % of built-in KAP meter readings were noted. Energy-dependent calibration factors are needed for unbiased PKA Accurate KAP meter calibration as a prerequisite for optimisation in projection radiography. PMID:26743261

  8. Active point out-of-plane ultrasound calibration

    Science.gov (United States)

    Cheng, Alexis; Guo, Xiaoyu; Zhang, Haichong K.; Kang, Hyunjae; Etienne-Cummings, Ralph; Boctor, Emad M.

    2015-03-01

    Image-guided surgery systems are often used to provide surgeons with informational support. Due to several unique advantages such as ease of use, real-time image acquisition, and no ionizing radiation, ultrasound is a common intraoperative medical imaging modality used in image-guided surgery systems. To perform advanced forms of guidance with ultrasound, such as virtual image overlays or automated robotic actuation, an ultrasound calibration process must be performed. This process recovers the rigid body transformation between a tracked marker attached to the transducer and the ultrasound image. Point-based phantoms are considered to be accurate, but their calibration framework assumes that the point is in the image plane. In this work, we present the use of an active point phantom and a calibration framework that accounts for the elevational uncertainty of the point. Given the lateral and axial position of the point in the ultrasound image, we approximate a circle in the axial-elevational plane with a radius equal to the axial position. The standard approach transforms all of the imaged points to be a single physical point. In our approach, we minimize the distances between the circular subsets of each image, with them ideally intersecting at a single point. We simulated in noiseless and noisy cases, presenting results on out-of-plane estimation errors, calibration estimation errors, and point reconstruction precision. We also performed an experiment using a robot arm as the tracker, resulting in a point reconstruction precision of 0.64mm.

  9. White Dwarfs for Calibrating the Dark Energy Survey

    Science.gov (United States)

    Allyn Smith, J.; Wester, William; Tucker, Douglas Lee; Fix, Mees B.; Tremblay, Pier-Emmanuel; Gulledge, Deborah J.; McDonald, Christopher P.; Allam, Sahar S.; James, David

    2016-01-01

    The Dark Energy Survey (DES) is surveying some 5000 square degrees in the southern hemisphere in the grizY filter system using the new Dark Energy Camera. In order to verify meeting photometric calibration requirements, we are obtaining imaging of several hundred white dwarfs (confirmed and candidates) to select nearly 100 or more hydrogen atmosphere (DA) white dwarfs for spectroscopy in the DES footprint. The spectra that are obtained will be extracted and used to derive synthetic spectra that can be compared with DES measurements from imaging in each of the DES grizY filters. This comparison should be able to verify and help calibrate the survey to a level better than 2% photometrically and to better than 0.5% in colors. We will discuss the observational and modeling effort required to develop a well-characterized DAs sample and present some preliminary results. This set would form the basis of a larger set of southern hemisphere survey calibration stars, and additionally serve as a legacy calibration set in the upcoming era of the LSST survey and the giant segmented mirror observatories. These stars will be used to establish and monitor the color zero points for the DES photometric system and can be used to search for systematic errors in the color zero points over the DES footprint. These stars will also be used as some of the primary standards for the DES photometric system which will allow nightly atmospheric monitoring during DES operations.

  10. Calibration of TOB+ Thermometer's Cards

    CERN Document Server

    Banitt, Daniel

    2014-01-01

    Motivation - Under the new upgrade of the CMS detector the working temperature of the trackers had been reduced to -27 Celsius degrees. Though the thermal sensors themselves (Murata and Fenwal thermistors) are effective at these temperatures, the max1542 PLC (programmable logic controller) cards, interpreting the resistance of the thermal sensors into DC counts usable by the DCS (detector control system), are not designed for these temperatures in which the counts exceed their saturation and therefor had to be replaced. In my project I was in charge of handling the emplacement and calibration of the new PLC cards to the TOB (tracker outer barrel) control system.

  11. Pulse-based internal calibration of polarimetric SAR

    DEFF Research Database (Denmark)

    Dall, Jørgen; Skou, Niels; Christensen, Erik Lintz

    1994-01-01

    Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops and devel......Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops...

  12. Tracked ultrasound calibration studies with a phantom made of LEGO bricks

    Science.gov (United States)

    Soehl, Marie; Walsh, Ryan; Rankin, Adam; Lasso, Andras; Fichtinger, Gabor

    2014-03-01

    In this study, spatial calibration of tracked ultrasound was compared by using a calibration phantom made of LEGO® bricks and two 3-D printed N-wire phantoms. METHODS: The accuracy and variance of calibrations were compared under a variety of operating conditions. Twenty trials were performed using an electromagnetic tracking device with a linear probe and three trials were performed using varied probes, varied tracking devices and the three aforementioned phantoms. The accuracy and variance of spatial calibrations found through the standard deviation and error of the 3-D image reprojection were used to compare the calibrations produced from the phantoms. RESULTS: This study found no significant difference between the measured variables of the calibrations. The average standard deviation of multiple 3-D image reprojections with the highest performing printed phantom and those from the phantom made of LEGO® bricks differed by 0.05 mm and the error of the reprojections differed by 0.13 mm. CONCLUSION: Given that the phantom made of LEGO® bricks is significantly less expensive, more readily available, and more easily modified than precision-machined N-wire phantoms, it prompts to be a viable calibration tool especially for quick laboratory research and proof of concept implementations of tracked ultrasound navigation.

  13. Calibrations and evaluation of the quality assurance during 1999 at the National Laboratory for ionising radiation

    International Nuclear Information System (INIS)

    The Swedish Radiation Protection Institute is the National Laboratory for the dosimetric quantities kerma, absorbed dose and dose equivalent. The activity is based on established calibration procedures and a quality assurance program for the used standards. This report gives a brief summary of the calibrations performed during 1999 and a more detailed description and analysis of the quality assurance during this year. The report makes it easier to draw conclusions about the long-term stability and possible malfunctions

  14. Sand Dune Ridge Alignment Effects on Surface BRF over the Libya-4 CEOS Calibration Site

    OpenAIRE

    Govaerts, Yves M.

    2015-01-01

    The Libya-4 desert area, located in the Great Sand Sea, is one of the most important bright desert CEOS pseudo-invariant calibration sites by its size and radiometric stability. This site is intensively used for radiometer drift monitoring, sensor intercalibration and as an absolute calibration reference based on simulated radiances traceable to the SI standard. The Libya-4 morphology is composed of oriented sand dunes shaped by dominant winds. The effects of sand dune spatial organization on...

  15. Assessment of Radiographic Image Quality by Visual Examination of Neutron Radiographs of the Calibration Fuel Pin

    OpenAIRE

    Domanus, Joseph Czeslaw

    1986-01-01

    Up till now no reliable radiographic image quality standards exist for neutron radiography of nuclear reactor fuel. Under the Euratoro Neutron Radiography Working Group (NRWG) Test Program neutron radiographs were produced at different neutron radiography facilities within the European Community of a calibration fuel pin. The radiographs were made by the direct, transfer and tracketch methods using different film recording materials. These neutron radiographs of the calibration fuel pin were ...

  16. In Search of Easy-to-Use Methods for Calibrating ADCP's for Velocity and Discharge Measurements

    Science.gov (United States)

    Oberg, K.; ,

    2002-01-01

    A cost-effective procedure for calibrating acoustic Doppler current profilers (ADCP) in the field was presented. The advantages and disadvantages of various methods which are used for calibrating ADCP were discussed. The proposed method requires the use of differential global positioning system (DGPS) with sub-meter accuracy and standard software for collecting ADCP data. The method involves traversing a long (400-800 meter) course at a constant compass heading and speed, while collecting simultaneous DGPS and ADCP data.

  17. An investigation of HPGe gamma efficiency calibration software (ANGLE V.3) for applications in nuclear decommissioning.

    Science.gov (United States)

    Bell, S J; Judge, S M; Regan, P H

    2012-12-01

    High resolution gamma spectrometry offers a rapid method to characterise waste materials on a decommissioning nuclear site. To meet regulatory requirements, measurements must be traceable to national standards, meaning that the spectrometers must be calibrated for a wide range of materials. Semi-empirical modelling software (such as ANGLE™) offers a convenient method to carry out such calibrations. This paper describes an assessment of the modelling software for use by a small laboratory based on a nuclear site. The results confirmed the need for accurate information on the detection construction if the calibration were to be accurate to within 10%. PMID:23041778

  18. IFIN-HH ionization chamber calibration and its validation; electrometric system improvement

    Energy Technology Data Exchange (ETDEWEB)

    Sahagia, M., E-mail: msahagia@nipne.r [' Horia Hulubei' National Institute of R and D for Physics and Nuclear Engineering, IFIN-HH, P.O. Box MG-6, Bucharest, RO-77125 (Romania); Waetjen, A.C.; Luca, A.; Ivan, C. [' Horia Hulubei' National Institute of R and D for Physics and Nuclear Engineering, IFIN-HH, P.O. Box MG-6, Bucharest, RO-77125 (Romania)

    2010-07-15

    The paper presents the results obtained in the calibration of the CENTRONIC IG12/20A ionization chamber for 18 gamma ray emitters, and its improvement with a Keithley 6517A electrometer. The calibration figures were determined either directly in pA MBq{sup -1} units, or calculated from old units, by using the system capacity value. The calibration figures, determined with RML's standards, are compared with those deduced from the KCRV or the mean of the comparisons, and the values determined at PTB-Germany.

  19. Calibration and efficiency curve of SANAEM ionization chamber for activity measurements.

    Science.gov (United States)

    Yeltepe, Emin; Kossert, Karsten; Dirican, Abdullah; Nähle, Ole; Niedergesäß, Christiane; Kemal Şahin, Namik

    2016-03-01

    A commercially available Fidelis ionization chamber was calibrated and assessed in PTB with activity standard solutions. The long-term stability and linearity of the system was checked. Energy-dependent efficiency curves for photons and beta particles were determined, using an iterative method in Excel™, to enable calibration factors to be calculated for radionuclides which were not used in the calibration. Relative deviations between experimental and calculated radionuclide efficiencies are of the order of 1% for most photon emitters and below 5% for pure beta emitters. The system will enable TAEK-SANAEM to provide traceable activity measurements.

  20. IFIN-HH ionization chamber calibration and its validation; electrometric system improvement.

    Science.gov (United States)

    Sahagia, M; Wätjen, A C; Luca, A; Ivan, C

    2010-01-01

    The paper presents the results obtained in the calibration of the CENTRONIC IG12/20A ionization chamber for 18 gamma ray emitters, and its improvement with a Keithley 6517A electrometer. The calibration figures were determined either directly in pA MBq(-1) units, or calculated from old units, by using the system capacity value. The calibration figures, determined with RML's standards, are compared with those deduced from the KCRV or the mean of the comparisons, and the values determined at PTB-Germany.