WorldWideScience

Sample records for calibration standards

  1. Radiological Calibration and Standards Facility

    Data.gov (United States)

    Federal Laboratory Consortium — PNNL maintains a state-of-the-art Radiological Calibration and Standards Laboratory on the Hanford Site at Richland, Washington. Laboratory staff provide expertise...

  2. Performance standard for dose Calibrator

    CERN Document Server

    Darmawati, S

    2002-01-01

    Dose calibrator is an instrument used in hospitals to determine the activity of radionuclide for nuclear medicine purposes. International Electrotechnical Commission (IEC) has published IEC 1303:1994 standard that can be used as guidance to test the performance of the instrument. This paper briefly describes content of the document,as well as explains the assessment that had been carried out to test the instrument accuracy in Indonesia through intercomparison measurement.Its is suggested that hospitals acquire a medical physicist to perform the test for its dose calibrator. The need for performance standard in the form of Indonesia Standard is also touched.

  3. AFFTC Standard Airspeed Calibration Procedures

    Science.gov (United States)

    1981-06-01

    25x0UIXQXQ Results of groundLpeed course calibration are normally pre- sented in the following plots: 1. .AvP vs Vi Ŗ. All vs V ic 3. AMPC vs Mic .4...8217Average AfPeavgpo, tion correction AM /AH 10-5 per and figure V 9 PC PC feet . fu V AYpc" x q3 @ , Average position avg corred ion (AM @ AMPC /AVPC...instrument error 0 M ic From and 0), Mach number p Chart 8.5 in reference’l (AFTR 6273) (DO AMPPacer poqition error calibra- Pc tion at9 S( AMpc /’,HpC)p

  4. Intercomparison of Laboratory Radiance Calibration Standards

    Science.gov (United States)

    Pavri, Betina; Chrien, Tom; Green, Robert; Williams, Orlesa

    2000-01-01

    Several standards for radiometric calibration were measured repeatedly with a spectroradiometer in order to understand how they compared in accuracy and stability. The tested radiance standards included a NIST 1000 W bulb and halon panel, two calibrated and stabilized integrating spheres, and a cavity blackbody. Results indicate good agreement between the blackbody and 1000 W bulb/spectralon panel, If these two radiance sources are assumed correct, then the integrating spheres did not conform. to their manufacturer-reported radiances in several regions of the spectrum. More detailed measurements am underway to investigate the discrepancy.

  5. 42 CFR 493.1255 - Standard: Calibration and calibration verification procedures.

    Science.gov (United States)

    2010-10-01

    ..., if possible, traceable to a reference method or reference material of known value; and (ii) Including... 42 Public Health 5 2010-10-01 2010-10-01 false Standard: Calibration and calibration verification... for Nonwaived Testing Analytic Systems § 493.1255 Standard: Calibration and calibration...

  6. Spectrophotometer spectral bandwidth calibration with absorption bands crystal standard.

    Science.gov (United States)

    Soares, O D; Costa, J L

    1999-04-01

    A procedure for calibration of a spectral bandwidth standard for high-resolution spectrophotometers is described. Symmetrical absorption bands for a crystal standard are adopted. The method relies on spectral band shape fitting followed by a convolution with the slit function of the spectrophotometer. A reference spectrophotometer is used to calibrate the spectral bandwidth standard. Bandwidth calibration curves for a minimum spectral transmission factor relative to the spectral bandwidth of the reference spectrophotometer are derived for the absorption bands at the wavelength of the band absorption maximum. The family of these calibration curves characterizes the spectral bandwidth standard. We calibrate the spectral bandwidth of a spectrophotometer with respect to the reference spectrophotometer by determining the spectral transmission factor minimum at every calibrated absorption band of the bandwidth standard for the nominal instrument values of the spectral bandwidth. With reference to the standard spectral bandwidth calibration curves, the relation of the spectral bandwidth to the reference spectrophotometer is determined. We determine the discrepancy in the spectrophotometers' spectral bandwidths by averaging the spectral bandwidth discrepancies relative to the standard calibrated values found at the absorption bands considered. A weighted average of the uncertainties is taken.

  7. Radioactive standards and calibration methods for contamination monitoring instruments

    Energy Technology Data Exchange (ETDEWEB)

    Yoshida, Makoto [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    1997-06-01

    Contamination monitoring in the facilities for handling unsealed radioactive materials is one of the most important procedures for radiation protection as well as radiation dose monitoring. For implementation of the proper contamination monitoring, radiation measuring instruments should not only be suitable to the purpose of monitoring, but also be well calibrated for the objective qualities of measurement. In the calibration of contamination monitoring instruments, quality reference activities need to be used. They are supplied in different such as extended sources, radioactive solutions or radioactive gases. These reference activities must be traceable to the national standards or equivalent standards. On the other hand, the appropriate calibration methods must be applied for each type of contamination monitoring instruments. In this paper, the concepts of calibration for contamination monitoring instruments, reference sources, determination methods of reference quantities and practical calibration methods of contamination monitoring instruments, including the procedures carried out in Japan Atomic Energy Research Institute and some relevant experimental data. (G.K.)

  8. Radiometric Calibrations, Measurements, and Standards Development at NREL: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Myers, D. R.; Andreas, A.; Stoffel, T.; Reda, I.; Wilcox, S.; Gotseff, P.; Kay, B.

    2001-10-01

    Presented at the 2001 NCPV Program Review Meeting: Radiometric calibrations, measurements, and standards development at NREL. We describe proposed revisions to current reference standard spectral distributions used to evaluate photovoltaic device performance and durability of materials. Improvements in broadband outdoor radiometer calibrations reduce uncertainties in broadband radiometer calibrations. We report a method to quantify the rate of change of broadband radiometer responsivities as a function of integrated exposure to irradiance and thermal energy. The results of applying a vector of calibration factors or responsivities to field data to remove zenith-angle dependent errors in global solar radiation measurements are shown. We report on the relative sensitivity of radiometers to daily versus biweekly cleaning.

  9. Syringe calibration factors and volume correction factors for the NPL secondary standard radionuclide calibrator

    CERN Document Server

    Tyler, D K

    2002-01-01

    The activity assay of a radiopharmaceutical administration to a patient is normally achieved via the use of a radionuclide calibrator. Because of the different geometries and elemental compositions between plastic syringes and glass vials, the calibration factors for syringes may well be significantly different from those for the glass containers. The magnitude of these differences depends on the energies of the emitted photons. For some radionuclides variations have been observed of 70 %, it is therefore important to recalibrate for syringes or use syringe calibration factors. Calibration factors and volume correction factors have been derived for the NPL secondary standard radionuclide calibrator, for a variety of commonly used syringes and needles, for the most commonly used medical radionuclide.

  10. NREL Spectral Standards Development and Broadband Radiometric Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Myers, D. R.; Andreas, A.; Stoffel, T.; Reda, I.; Wilcox, S.; Gotseff, P.; Kay, B.; Gueymard, C.

    2003-05-01

    We describe a final version of revisions to current ASTM reference standard spectral distributions used to evaluate photovoltaic device performance. An NREL-developed graphical user interface for working with the SMARTS2 spectral model has been developed and is being tested. A proposed ASTM reference Ultraviolet (UV) spectra for materials durability is presented. Improvements in broadband outdoor radiometer calibration, characterization, and reporting software reduce uncertainties in broadband radiometer calibrations.

  11. Standard guide for calibrating reticles and light microscope magnifications

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2002-01-01

    1.1 This guide covers methods for calculating and calibrating microscope magnifications, photographic magnifications, video monitor magnifications, grain size comparison reticles, and other measuring reticles. Reflected light microscopes are used to characterize material microstructures. Many materials engineering decisions may be based on qualitative and quantitative analyses of a microstructure. It is essential that microscope magnifications and reticle dimensions be accurate. 1.2 The calibration using these methods is only as precise as the measuring devices used. It is recommended that the stage micrometer or scale used in the calibration should be traceable to the National Institute of Standards and Technology (NIST) or a similar organization. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory lim...

  12. DNA origami structures as calibration standards for nanometrology

    Science.gov (United States)

    Korpelainen, Virpi; Linko, Veikko; Seppä, Jeremias; Lassila, Antti; Kostiainen, Mauri A.

    2017-03-01

    In this work we have studied the feasibility of DNA origami nanostructures as dimensional calibration standards for atomic force microscopes (AFMs) at the nanometre scale. The stability of the structures and repeatability of the measurement have been studied, and the applicability for calibration is discussed. A cross-like Seeman tile (ST) was selected for the studies and it was found suitable for repeatable calibration of AFMs. The height of the first height step of the ST was 2.0 nm. Expanded standard uncertainty (k  =  2) of the measurement U c was 0.2 nm. The width of the ST was 88 nm and width of its arm was 28 nm with U c   =  3 nm. In addition, prepared dry samples were found out to be stable at least for 12 months.

  13. Standard practice for torque calibration of testing machines and devices

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2009-01-01

    1.1 This practice covers procedures and requirements for the calibration of torque for static and quasi-static torque capable testing machines or devices. These may, or may not, have torque indicating systems and include those devices used for the calibration of hand torque tools. Testing machines may be calibrated by one of the three following methods or combination thereof: 1.1.1 Use of standard weights and lever arms. 1.1.2 Use of elastic torque measuring devices. 1.1.3 Use of elastic force measuring devices and lever arms. 1.1.4 Any of the methods require a specific uncertainty of measurement and a traceability derived from national standards of mass and length. 1.2 The procedures of 1.1.1, 1.1.2, and 1.1.3 apply to the calibration of the torque-indicating systems associated with the testing machine, such as a scale, dial, marked or unmarked recorder chart, digital display, etc. In all cases the buyer/owner/user must designate the torque-indicating system(s) to be calibrated and included in the repor...

  14. On Calibrations Using the Crab Nebula as a Standard Candle

    Science.gov (United States)

    Weisskopf, Martin; Guainazzi, Matteo; Jahoda, Keith; Shaposhnikov, Nikolai; ODell, Stephen; Zavlin, Vyacheslav; Wilson-Hodge, Colleen; Elsner, Ronald

    2009-01-01

    Inspired by a recent paper (Kirsch et al. 2005) on possible use of the Crab Nebula as a standard candle for calibrating X-ray response func tions, we examine possible consequences of intrinsic departures from a single (absorbed) power law upon such calibrations. We limited our analyses to three more modern X-ray instruments -- the ROSAT/PSPC, th e RXTE/PCA, and the XMM-Newton/EPIC-pn. The results are unexpected an d indicate a need to refine two of the three response functions studi ed. The implications for Chandra will be discussed.

  15. Monochromator-Based Absolute Calibration of a Standard Radiation Thermometer

    Science.gov (United States)

    Mantilla, J. M.; Hernanz, M. L.; Campos, J.; Martín, M. J.; Pons, A.; del Campo, D.

    2014-04-01

    Centro Español de Metrología (CEM) is disseminating the International Temperature Scale (ITS-90), at high temperatures, by using the fixed points of Ag and Cu and a standard radiation thermometer. However, the future mise-en-pratique for the definition of the kelvin ( MeP-K) will include the dissemination of the kelvin by primary methods and by indirect approximations capable of exceptionally low uncertainties or increased reliability. Primary radiometry is, at present, able to achieve uncertainties competitive with the ITS-90 above the silver point with one of the possible techniques the calibration for radiance responsivity of an imaging radiometer (radiance method). In order to carry out this calibration, IO-CSIC (Spanish Designated Institute for luminous intensity and luminous flux) has collaborated with CEM, allowing traceability to its cryogenic radiometer. A monochromator integrating sphere-based spectral comparator facility has been used to calibrate one of the CEM standard radiation thermometers. The absolute calibrated standard radiation thermometer has been used to determine the temperatures of the fixed points of Cu, Co-C, Pt-C, and Re-C. The results obtained are 1357.80 K, 1597.10 K, 2011.66 K, and 2747.64 K, respectively, with uncertainties ranging from 0.4 K to 1.1 K.

  16. A standard stellar library for evolutionary synthesis. III. Metallicity calibration

    Science.gov (United States)

    Westera, P.; Lejeune, T.; Buser, R.; Cuisinier, F.; Bruzual, G.

    2002-01-01

    We extend the colour calibration of the widely used BaSeL standard stellar library (Lejeune et al. 1997, 1998) to non-solar metallicities, down to [Fe/H] ~ -2.0 dex. Surprisingly, we find that at the present epoch it is virtually impossible to establish a unique calibration of UBVRIJHKL colours in terms of stellar metallicity [Fe/H] which is consistent simultaneously with both colour-temperature relations and colour-absolute magnitude diagrams (CMDs) based on observed globular cluster photometry data and on published, currently popular standard stellar evolutionary tracks and isochrones. The problem appears to be related to the long-standing incompleteness in our understanding of convection in late-type stellar evolution, but is also due to a serious lack of relevant observational calibration data that would help resolve, or at least further significant progress towards resolving this issue. In view of the most important applications of the BaSeL library, we here propose two different metallicity calibration versions: (1) the ``WLBC 99'' library, which consistently matches empirical colour-temperature relations and which, therefore, should make an ideal tool for the study of individual stars; and (2), the ``PADOVA 2000'' library, which provides isochrones from the Padova 2000 grid (Girardi et al. \\cite{padova}) that successfully reproduce Galactic globular-cluster colour-absolute magnitude diagrams and which thus should prove particularly useful for studies of collective phenomena in stellar populations in clusters and galaxies.

  17. Calibration of the Standards and Calibration Laboratory`s Co{sup 60} Radiation Pool

    Energy Technology Data Exchange (ETDEWEB)

    Wirtenson, G.R.; White, R.H.

    1993-01-01

    The authors report measurements of dose rates at various locations in the LLNL Standards and Calibrations Laboratory`s Co{sup 60} Radiation Pool. Plots show the dependence of dose rate on radius near the bottom of the pool and the dependence of dose rate on height at a fixed distance from the pool center. The effect of varying sample location within the pool`s dry-well was also investigated.

  18. Development of a low-level Ar-37 calibration standard

    CERN Document Server

    Williams, R M; Bowyer, T W; Day, A R; Fuller, E S; Haas, D A; Hayes, J C; Hoppe, E W; Humble, P H; Keillor, M E; LaFerriere, B D; Mace, E K; McIntyre, J I; Miley, H S; Myers, A W; Orrell, J L; Overman, C T; Panisko, M E; Seifert, A

    2016-01-01

    Argon-37 is an environmental signature of an underground nuclear explosion. Producing and quantifying low-level Ar-37 standards is an important step in the development of sensitive field measurement instruments. This paper describes progress at Pacific Northwest National Laboratory in developing a process to generate and quantify low-level Ar-37 standards, which can be used to calibrate sensitive field systems at activities consistent with soil background levels. This paper presents a discussion of the measurement analysis, along with assumptions and uncertainty estimates.

  19. Performance of calibration standards for antigen quantitation with flow cytometry.

    Science.gov (United States)

    Lenkei, R; Gratama, J W; Rothe, G; Schmitz, G; D'hautcourt, J L; Arekrans, A; Mandy, F; Marti, G

    1998-10-01

    In the frame of the activities initiated by the Task Force for Antigen Quantitation of the European Working Group on Clinical Cell Analysis (EWGCCA), an experiment was conducted to evaluate microbead standards used for quantitative flow cytometry (QFCM). An unified window of analysis (UWA) was established on three different instruments (EPICS XL [Coulter Corporation, Miami, FL], FACScan and FACS Calibur [Becton Dickinson, San Jose, CA]) with QC3 microbeads (FCSC, PR). By using this defined fluorescence intensity scale, the performance of several monoclonal antibodies directed to CD3, CD4, and CD8 (conjugated and unconjugated), from three manufacturers (BDIS, Coulter [Immunotech], and DAKO) was tested. In addition, the QIFI system (DAKO) and QuantiBRITE (BDIS), and a method of relative fluorescence intensity (RFI, method of Giorgi), were compared. mAbs reacting with three more antigens, CD16, CD19, and CD38 were tested on the FACScan instrument. Quantitation was carried out using a single batch of cryopreserved peripheral blood leukocytes, and all tests were performed as single color analyses. Significant correlations were observed between the antibody-binding capacity (ABC) values of the same CD antigen measured with various calibrators and with antibodies differing in respect to vendor, labeling and possible epitope recognition. Despite the significant correlations, the ABC values of most monoclonal antibodies differed by 20-40% when determined by the different fluorochrome conjugates and different calibrators. The results of this study indicate that, at the present stage of QFCM consistent ABC values may be attained between laboratories provided that a specific calibration system is used including specific calibrators, reagents, and protocols.

  20. A report from the AVS Standards Committee - Comparison of ion gauge calibrations by several standards laboratories

    Science.gov (United States)

    Warshawsky, I.

    1982-01-01

    Calibrations by four U.S. laboratories of four hot-cathode ion gauges, in the range 0.07-13 mPa, showed systematic differences among laboratories that were much larger than the expected error of any one calibration. They also suggested that any of the four gauges tested, if properly packaged and shipped, was able to serve as a transfer standard with probable error of 2%. A second comparison was made of the calibrations by two U.S. laboratories of some other gauges that had also been calibrated by the National Physical Laboratory, England. Results did not permit conclusive determination of whether differences were due to the laboratories or to changes in the gauges.

  1. Internal standard versus external standard calibration: an uncertainty case study of a liquid chromatography analysis

    Directory of Open Access Journals (Sweden)

    Elcio Cruz de Oliveira

    2010-01-01

    Full Text Available Traditionally, in the cigarettes industry, the determination of ammonium ion in the mainstream smoke is performed by ion chromatography. This work studies this determination and compares the results of this technique with the use of external and internal standard calibration. A reference cigarette sample presented measurement uncertainty of 2.0 μg/cigarette and 1.5 μg/cigarette, with external and internal standard, respectively. It is observed that the greatest source of uncertainty is the bias correction factor and that it is even more significant when using external standard, confirming thus the importance of internal standardization for this correction.

  2. Calibration Standards for Surface Topography Measuring Systems down to Nanometric Range

    DEFF Research Database (Denmark)

    Trumpold, H.; De Chiffre, Leonardo; Andreasen, Jan Lasson

    calibrator has been designed, built and tested. Primary calibration standards have been produced by ion-beam and plasma etching (step height standards), by holographic generation of sinusoidal structures with two-beam interference exposure and by ultra-precision diamond cutting. From primary standards...... to be replicated during all stages of the replication processes. Procedures for cleaning glass, PVC, PC, PMM and Ni-surfaces have been developed and tested. Calibration procedures for calibration standards and for calibrating instruments in X-, Y- and Z-direction have been developed and tested. Proposals...

  3. STANDARD STARS AND EMPIRICAL CALIBRATIONS FOR Hα AND Hβ PHOTOMETRY

    Energy Technology Data Exchange (ETDEWEB)

    Joner, Michael D.; Hintz, Eric G., E-mail: joner@byu.edu, E-mail: hintz@byu.edu [Department of Physics and Astronomy, Brigham Young University, N283 ESC, Provo, UT 84602 (United States)

    2015-12-15

    We define an Hα photometric system that is designed as a companion to the well established Hβ index. The new system is built on spectrophotometric observations of field stars as well as stars in benchmark open clusters. We present data for 75 field stars, 12 stars from the Coma star cluster, 24 stars from the Hyades, 17 stars from the Pleiades, and 8 stars from NGC 752 to be used as primary standard stars in the new systems. We show that the system transformations are relatively insensitive to the shape of the filter functions. We make comparisons of the Hα index to the Hβ index and illustrate the relationship between the two systems. In addition, we present relations that relate both hydrogen indices to equivalent width and effective temperature. We derive equations to calibrate both systems for Main Sequence stars with spectral types in the range O9 to K2 for equivalent width and A2 to K2 for effective temperature.

  4. Standard guide for establishing calibration for a measurement method used to analyze nuclear fuel cycle materials

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2003-01-01

    1.1 This guide provides the basis for establishing calibration for a measurement method typically used in an analytical chemistry laboratory analyzing nuclear materials. Guidance is included for such activities as preparing a calibration procedure, selecting a calibration standard, controlling calibrated equipment, and documenting calibration. The guide is generic and any required technical information specific for a given method must be obtained from other sources.

  5. Electronic Correlated Noise Calibration Standard for Interferometric and Polarimetric Microwave Radiometers Project

    Data.gov (United States)

    National Aeronautics and Space Administration — A new type of calibration standard is proposed which produces a pair of microwave noise signals to aid in the characterization and calibration of correlating...

  6. The Moon as a photometric calibration standard for microwave sensors

    Science.gov (United States)

    Burgdorf, Martin; Buehler, Stefan A.; Lang, Theresa; Michel, Simon; Hans, Imke

    2016-08-01

    Instruments on satellites for Earth observation on polar orbits usually employ a two-point calibration technique, in which deep space and an onboard calibration target provide two reference flux levels. As the direction of the deep-space view is in general close to the celestial equator, the Moon sometimes moves through the field of view and introduces an unwelcome additional signal. One can take advantage of this intrusion, however, by using the Moon as a third flux standard, and this has actually been done for checking the lifetime stability of sensors operating at visible wavelengths. As the disk-integrated thermal emission of the Moon is less well known than its reflected sunlight, this concept can in the microwave range only be used for stability checks and intercalibration. An estimate of the frequency of appearances of the Moon in the deep-space view, a description of the limiting factors of the measurement accuracy and models of the Moon's brightness, and a discussion of the benefits from complementing the naturally occurring appearances of the Moon with dedicated spacecraft maneuvers show that it would be possible to detect photometric lifetime drifts of a few percent with just two measurements. The pointing accuracy is the most crucial factor for the value of this method. Planning such observations in advance would be particularly beneficial, because it allows observing the Moon at well-defined phase angles and putting it at the center of the field of view. A constant phase angle eliminates the need for a model of the Moon's brightness when checking the stability of an instrument. With increasing spatial resolution of future microwave sensors another question arises, viz. to what extent foreground emission from objects other than the Moon will contaminate the flux entering the deep-space view, which is supposed to originate exclusively in the cosmic microwave background. We conclude that even the brightest discreet sources have flux densities below the

  7. Stability study of standards used for calibration of the spectrophotometer wavelength scale

    Science.gov (United States)

    Debossan, L. F.; Carvalho, E. M. S.; Souza, M. A.; Gomes, J. F. S.

    2016-07-01

    The calibration of spectrophotometers is a procedure recommended by international standards to provide quality assurance of results and traceability. Due to its intrinsic properties, holmium oxide filters are indicated as reference standards for calibrating the wavelength scale of such equipment. This paper presents a study aiming to assess the repeatability and drift of holmium oxide standard filters calibrated in the Radiometry and Photometry Laboratory (Laraf) of Inmetro in order to verify their stability.

  8. Standard practice for alternate actinide calibration for inductively coupled plasma-mass spectrometry

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2004-01-01

    1.1 This practice provides guidance for an alternate linear calibration for the determination of selected actinide isotopes in appropriately prepared aqueous solutions by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS). This alternate calibration is mass bias adjusted using thorium-232 (232Th) and uranium-238 (238U) standards. One of the benefits of this standard practice is the ability to calibrate for the analysis of highly radioactive actinides using calibration standards at much lower specific activities. Environmental laboratories may find this standard practice useful if facilities are not available to handle the highly radioactive standards of the individual actinides of interest. 1.2 The instrument response for a series of determinations of known concentration of 232Th and 238U defines the mass versus response relationship. For each standard concentration, the slope of the line defined by 232Th and 238U is used to derive linear calibration curves for each mass of interest using interference equ...

  9. Development of a Low-Level Ar-37 Calibration Standard

    Energy Technology Data Exchange (ETDEWEB)

    Williams, Richard M.; Aalseth, Craig E.; Bowyer, Ted W.; Day, Anthony R.; Fuller, Erin S.; Haas, Derek A.; Hayes, James C.; Hoppe, Eric W.; Humble, Paul H.; Keillor, Martin E.; LaFerriere, Brian D.; Mace, Emily K.; McIntyre, Justin I.; Miley, Harry S.; Myers, Allan W.; Orrell, John L.; Overman, Cory T.; Panisko, Mark E.; Seifert, Allen

    2016-03-07

    Argon-37 is an important environmental signature of an underground nuclear explosion. Producing and quantifying low-level 37Ar standards is an important step in the development of sensitive field measurement instruments for use during an On-Site Inspection, a key provision of the Comprehensive Nuclear-Test-Ban Treaty. This paper describes progress at Pacific Northwest National Laboratory (PNNL) in the development of a process to generate and quantify low-level 37Ar standard material, which can then be used to calibrate sensitive field systems at activities consistent with soil background levels. The 37Ar used for our work was generated using a laboratory-scale, high-energy neutron source to irradiate powdered samples of calcium carbonate. Small aliquots of 37Ar were then extracted from the head space of the irradiated samples. The specific activity of the head space samples, mixed with P10 (90% stable argon:10% methane by mole fraction) count gas, is then derived using the accepted Length-Compensated Internal-Source Proportional Counting method. Due to the low activity of the samples, a set of three Ultra-Low Background Proportional-Counters designed and fabricated at PNNL from radio-pure electroformed copper was used to make the measurements in PNNL’s shallow underground counting laboratory. Very low background levels (<10 counts/day) have been observed in the spectral region near the 37Ar emission feature at 2.8 keV. Two separate samples from the same irradiation were measured. The first sample was counted for 12 days beginning 28 days after irradiation, the second sample was counted for 24 days beginning 70 days after irradiation (the half-life of 37Ar is 35.0 days). Both sets of measurements were analyzed and yielded very similar results for the starting activity (~0.1 Bq) and activity concentration (0.15 mBq/ccSTP argon) after P10 count gas was added. A detailed uncertainty model was developed based on the ISO Guide to the Expression of Uncertainty in

  10. Standard Procedure for Calibrating an Areal Calorimetry Based Dosimeter

    Science.gov (United States)

    2015-05-01

    unlimited (P.A. Case No. TSRL-PA-2015-0136, 24 Sep 15). REFERENCES Beason, C. W. (2005). CLT As A Calorimeter. PowerPoint presentation dated 10 January...or sell any patented invention that may relate to them. This report was cleared for public release by the 88th ABW Public Affairs Office and is...is presented for calibrating an areal calorimetry-based dosimeter. A detailed example of calibrating an areal calorimetry-based dosimeter is

  11. Lyophilized standards for the calibration of real time PCR assay for hepatitis C virus RNA

    Institute of Scientific and Technical Information of China (English)

    WANG Lu-nan; WU Jian-min; DENG Wei; SHEN Zi-yu; CHEN Wen-xiang; LI Jin-ming

    2006-01-01

    Background Since October 1997, an international standard for hepatitis C virus (HCV) nucleic acid amplification technology assay, 96/790, has been available. We compared a series of lyophilized standards with known HCV RNA concentrations against the international standard in fluorescence quantitative PCR detection.Methods A series of lyophilized sera were calibrated by ROCHE COBAS AMPLICOR HCV Monitor test against the international standard and sent to various manufacturers to analyse the samples using their own kits.Then calibration curves from the series were compared with that obtained from the external standard calibration curve with the manufacture's series.Results The standard calibration curve with the series of lyophilized serum showed an excellent correlation(R2>0.98), slope and intercept that were similar to those from the manufacture's series. When the standard calibration curve from the series of lyophilized standards were used to define the values of the given sample,lower coefficients of variation between kits from different manufactures were obtained.Conclusion The results showed that the lyophilized standards could be used to setup the standard calibration curve for clinical HCV RNA quantitative PCR detection.

  12. Source geometry factors for HDR 192Ir brachytherapy secondary standard well-type ionization chamber calibrations

    Science.gov (United States)

    Shipley, D. R.; Sander, T.; Nutbrown, R. F.

    2015-03-01

    Well-type ionization chambers are used for measuring the source strength of radioactive brachytherapy sources before clinical use. Initially, the well chambers are calibrated against a suitable national standard. For high dose rate (HDR) 192Ir, this calibration is usually a two-step process. Firstly, the calibration source is traceably calibrated against an air kerma primary standard in terms of either reference air kerma rate or air kerma strength. The calibrated 192Ir source is then used to calibrate the secondary standard well-type ionization chamber. Calibration laboratories are usually only equipped with one type of HDR 192Ir source. If the clinical source type is different from that used for the calibration of the well chamber at the standards laboratory, a source geometry factor, ksg, is required to correct the calibration coefficient for any change of the well chamber response due to geometric differences between the sources. In this work we present source geometry factors for six different HDR 192Ir brachytherapy sources which have been determined using Monte Carlo techniques for a specific ionization chamber, the Standard Imaging HDR 1000 Plus well chamber with a type 70010 HDR iridium source holder. The calculated correction factors were normalized to the old and new type of calibration source used at the National Physical Laboratory. With the old Nucletron microSelectron-v1 (classic) HDR 192Ir calibration source, ksg was found to be in the range 0.983 to 0.999 and with the new Isodose Control HDR 192Ir Flexisource ksg was found to be in the range 0.987 to 1.004 with a relative uncertainty of 0.4% (k = 2). Source geometry factors for different combinations of calibration sources, clinical sources, well chambers and associated source holders, can be calculated with the formalism discussed in this paper.

  13. Improved Evaluation of Planar Calibration Standards Using the TDR Preselection Method

    Directory of Open Access Journals (Sweden)

    J. Vancl

    2007-01-01

    Full Text Available Calibration and correction methods for the Vector Network Analyzer (VNA are based on the fundamental assumption of the constant error model, which is independent of connected calibration standards and/or devices under test (DUT. Unfortunately, this assumption is not satisfied well for planar calibration standards fabricated by etching technology on soft substrates. An evaluation of the error model is affected especially by variations in the manufacturing process and also by the reproducibility of an assembly. In this paper, we propose error minimization by selecting the best combination of available calibration standards based on time domain reflection (TDR measurement, which can also be obtained by the fourier transformation from the measured S-parameters. The proposed method was verified experimentally using short, open, load and thru (SOLT standards fabricated on an FR4 laminate substrate which achieves the essential reduction of the measurement error in the frequency range up to 15 GHz. 

  14. COMPARISON METHODOLOGIES FOR CALIBRATION OF Hp(10) PERSONAL DOSEMETERS USING ISO 4037 AND ISO 29661 STANDARDS.

    Science.gov (United States)

    Cardoso, J; Santos, L; Carvalhal, G; Oliveira, C

    2016-09-01

    The calibration of electronic personal dosemeters at the Portuguese ionizing radiation metrology laboratory uses the standard IEC 61526 for calibration methodology. This standard describes the irradiation geometry for testing and indicates that the standard ISO 4037-1, 2, 3 and 4 should be used. The ISO 4037 establishes that the reference point of test is a point in the radiation monitor, known or established, and the calibration phantom should be placed on its back in order to simulate the trunk body. Recently, ISO published another standard, the ISO 29661, that changes the reference point from the radiation monitor to the front face of the calibration phantom. The aim of this work is to present the result of the comparison of these two methodologies on personal dosemeters from five different manufacturers. The work shows differences in the Hp(10) response up to 4% resulting from the two different reference point concepts.

  15. Services of the CDRH X-ray calibration laboratory and their traceability to National Standards

    Energy Technology Data Exchange (ETDEWEB)

    Cerra, F.; Heaton, H.T. [Center for Devices and Radiological Health, Rockville, MD (United States)

    1993-12-31

    The X-ray Calibration Laboratory (XCL) of the Center for Devices and Radiological Health (CDRH) provides calibration services for the Food and Drug Administration (FDA). The instruments calibrated are used by FDA and contract state inspectors to verify compliance with federal x-ray performance standards and for national surveys of x-ray trends. In order to provide traceability of measurements, the CDRH XCL is accredited by the National Voluntary Laboratory Accreditation Program (NVLAP) for reference, diagnostic, and x-ray survey instrument calibrations. In addition to these accredited services, the CDRH XCL also calibrates non-invasive kVp meters in single- and three-phase x-ray beams, and thermoluminescent dosimeter (TLD) chips used to measure CT beam profiles. The poster illustrates these services and shows the traceability links back to the National Standards.

  16. Establishment of the platinum-iridium kilogram mass standards at NMIJ after the Extraordinary Calibrations

    Science.gov (United States)

    Mizushima, Shigeki; Fujii, Kenichi

    2016-04-01

    The Bureau International des Poids et Mesures has carried out calibrations of the platinum-iridium kilogram mass standards by referencing the international prototype of the kilogram for the first time since the third periodic verification of national prototypes of the kilogram was carried out in 1988-92. This calibration campaign was designated ‘Extraordinary Calibrations’ in the second phase, in which two platinum-iridium kilogram mass standards of the National Metrology Institute of Japan were calibrated with a standard uncertainty of 3.5 μg. By adding these new calibration data into our data sets from 1991, we established our mass unit with a standard uncertainty of 3.3 μg by least-squares analysis using an exponential model, which is useful for compensating for mass increase after cleaning the mass standards. Moreover, it was found that our established mass unit following the Extraordinary Calibrations shifted against our previously maintained mass unit by  -20.8 μg as of the beginning of 2015. The analysis with a linear model revealed that the amount of mass increase over time of some standards was significantly smaller than that suggested at the third periodic verification of national prototypes of the kilogram. The analysis with the exponential model gave an exponent of 0.217 with a standard uncertainty of 0.057. This suggests that the mass increase due to surface contamination cannot be explained by a diffusion-limited process.

  17. Calibration of the SH134-20 Standard Gain Horn

    DEFF Research Database (Denmark)

    Pivnenko, Sergey; Breinbjerg, Olav

    This report documents the measurement of the linearly polarized SH134-20 Standard Gain Horn. The measurement comprises on-axis gain, on-axis polarization characteristics, and reflection coefficient at 111 frequencies in the frequency range from 22-33 GHz. The measurement was carried out at the DT......-ESA Spherical Near-Field Antenna Test Facility in December 2012 for SE Laboratories, Santa Clara, CA, USA....

  18. Satellite-Mounted Light Sources as Photometric Calibration Standards

    CERN Document Server

    Albert, Justin; Battat, James; Dupuis, Grace; Fransham, Kyle; Koopmans, Kristin; Jarrett, Michael

    2009-01-01

    A significant and growing portion of systematic error on a number of fundamental parameters in astrophysics and cosmology is due to uncertainties from absolute photometric and flux standards. A path toward achieving major reduction in such uncertainties may be provided by satellite-mounted light sources, resulting in improvement in the ability to precisely characterize atmospheric extinction, and thus helping to usher in the coming generation of precision results in astronomy. Toward this end, we have performed a campaign of observations of the 532 nm pulsed laser aboard the CALIPSO satellite, using a portable network of cameras and photodiodes, to precisely measure atmospheric extinction.

  19. Quantitative Analysis of Carbon Steel with Multi-Line Internal Standard Calibration Method Using Laser-Induced Breakdown Spectroscopy.

    Science.gov (United States)

    Pan, Congyuan; Du, Xuewei; An, Ning; Zeng, Qiang; Wang, Shengbo; Wang, Qiuping

    2016-04-01

    A multi-line internal standard calibration method is proposed for the quantitative analysis of carbon steel using laser-induced breakdown spectroscopy (LIBS). A procedure based on the method was adopted to select the best calibration curves and the corresponding emission lines pairs automatically. Laser-induced breakdown spectroscopy experiments with carbon steel samples were performed, and C, Cr, and Mn were analyzed via the proposed method. Calibration curves of these elements were constructed via a traditional single line internal standard calibration method and a multi-line internal standard calibration method. The calibration curves obtained were evaluated with the determination coefficient, the root mean square error of cross-validation, and the average relative error of cross-validation. All of the parameters were improved significantly with the proposed method. The results show that accurate and stable calibration curves can be obtained efficiently via the multi-line internal standard calibration method.

  20. Investigation of Ultrasonic Calibration using Steel Standard Reference Blocks

    Directory of Open Access Journals (Sweden)

    Mohamed Galal Sayed ALI

    2012-12-01

    Full Text Available The equipment used in standardization must be adjusted in order that both the receiver amplifier and video display are linear over the range of distance and flaw size to be used. The failure to do that would result in different block-to-block response ratios because of the measurements depending on the scope reading or attenuator setting. In this paper, using two sets of steel reference blocks (area-amplitude and distance-amplitude to modified method to verify the linearity of the system of instruments used in the ultrasonic inspection. This method depends on the response from reference blocks with different hole diameters at constant gain and ratio between the figures slope at different gains. The linearity of the ultrasonic instruments can be measured with high precision. The results can be used effectively for determining resolution and sensitivity capabilities.

  1. Calibration and commutability assessment of the 1st International Standard for Diphtheria Antitoxin Human.

    Science.gov (United States)

    Stickings, Paul; Rigsby, Peter; Coombes, Laura; von Hunolstein, Christina; Ralli, Luisa; Pinto, Antonella; Sesardic, Dorothea

    2013-11-01

    The 1st International Standard for Diphtheria Antitoxin Human (coded 10/262) was established by the World Health Organization Expert Committee on Biological Standardization in 2012. This paper describes the production, characterization and calibration of the new standard which is intended for use in the standardization of assays used to measure diphtheria antibody responses in human serum. The new standard was calibrated in terms of the International Standard for Diphtheria Antitoxin Equine in an international collaborative study. A total of 8 participants from 8 different countries performed in vivo and/or in vitro toxin neutralization tests and returned data that was used to assign units to the proposed new standard. The new standard has a diphtheria antitoxin potency of 2 IU/ampoule and is predicted to be stable. A follow up study was performed to assess commutability of the new standard. The follow up study was an existing external quality assessment, modified to include the new standard. Results obtained suggest that the new standard is commutable, showing comparable behaviour to native human serum samples in the majority of the assays compared, and is therefore suitable for use as a reference preparation in assays used to measure the level of anti-diphtheria antibodies in human serum.

  2. Assessment of target volume doses in radiotherapy based on the standard and measured calibration curves

    Directory of Open Access Journals (Sweden)

    Gholamreza Fallah Mohammadi

    2015-01-01

    Full Text Available Context: In radiation treatments, estimation of the dose distribution in the target volume is one of the main components of the treatment planning procedure. To estimate the dose distribution, the information of electron densities is necessary. The standard curves determined by computed tomography (CT scanner that may be different from that of other oncology centers. In this study, the changes of dose calculation due to the different calibration curves (HU-ρel were investigated. Materials and Methods: Dose values were calculated based on the standard calibration curve that was predefined for the treatment planning system (TPS. The calibration curve was also extracted from the CT images of the phantom, and dose values were calculated based on this curve. The percentage errors of the calculated values were determined. Statistical Analysis Used: The statistical analyses of the mean differences were performed using the Wilcoxon rank-sum test for both of the calibration curves. Results and Discussion: The results show no significant difference for both of the measured and standard calibration curves (HU-ρel in 6, 15, and 18 MeV energies. In Wilcoxon ranked sum nonparametric test for independent samples with P < 0.05, the equality of monitor units for both of the curves to transfer 200 cGy doses to reference points was resulted. The percentage errors of the calculated values were lower than 2% and 1.5% in 6 and 15 MeV, respectively. Conclusion: From the results, it could be concluded that the standard calibration curve could be used in TPS dose calculation accurately.

  3. Transfer standard device to improve the traceable calibration of physiotherapy ultrasound machines

    NARCIS (Netherlands)

    Hekkenberg, R.T.; Richards, A.; Beissner, K.; Zeqiri, B.; Bezemer, R.A.; Hodnett, M.; Prout, G.; Cantrall, C.

    2006-01-01

    Ultrasound (US) physiotherapy as a clinical treatment is extremely common in the Western world. Internationally, regulation to ensure safe application of US physiotherapy by regular calibration ranges from nil to mandatory. The need for a portable power standard (PPS) has been addressed within a Eur

  4. Nickel replicas as calibration reference standards for industrial surface texture instruments

    DEFF Research Database (Denmark)

    Sammatini-Malberg, Maria-Pia

    The present report is a documentation of measurements carried out at DTU on Nickel replicas. The research is performed in the frame of the project with contract SMT4-CT97-2176 with title: Calibration Standards for Surface Topography Measuring Systems down to Nanometric Scale....

  5. Secondary calibration laboratory for ionizing radiation laboratory accreitation program National Institute of Standards and Technology National Voluntary Laboratory Accreditation Program

    Energy Technology Data Exchange (ETDEWEB)

    Martin, P.R.

    1993-12-31

    This paper presents an overview of the procedures and requirements for accreditation under the Secondary Calibration Laboratory for Ionizing Radiation Program (SCLIR LAP). The requirements for a quality system, proficiency testing and the onsite assessment are discussed. The purpose of the accreditation program is to establish a network of secondary calibration laboratories that can provide calibrations traceable to the primary national standards.

  6. SIMULTANEOUS CALIBRATION OF MOLECULAR WEIGHT SEPARATION AND COLUMN DISPERSION OF SEC WITH CHARACTERIZED POLYMER STANDARDS

    Institute of Scientific and Technical Information of China (English)

    CHENG Rongshi; BO Shuqin

    1983-01-01

    With the aid of the theoretical relationship between the calibration relation of a SEC column for the monodisperse polymer species under ideal working condition and the effective relations between the molecular weight and the elution volume for characterized polymer samples, a computational procedure for simultaneous calibration of molecular weight separation and column dispersion is proposed. From the experimental chromatograms of narrow MWD polystyrene standards and broad MWD 1,2-polybutadiene fractions the spreading factors of a SEC column was deduced by the proposed method. The variation of the spreading factor with the elution volume is independent upon the polymer sample used.

  7. One-Port Direct/Reverse Method for Characterizing VNA Calibration Standards

    CERN Document Server

    Monsalve, Raul A; Mozdzen, Thomas J; Bowman, Judd D

    2016-01-01

    This paper introduces a one-port method for estimating model parameters of VNA calibration standards. The method involves measuring the standards through an asymmetrical passive network connected in direct mode and then in reverse mode, and using these measurements to compute the S-parameters of the network. The free parameters of the calibration standards are estimated by minimizing a figure of merit based on the expected equality of the S-parameters of the network when used in direct and reverse modes. The capabilities of the method are demonstrated through simulations, and real measurements are used to estimate the actual offset delay of a 50-$\\mathbf{\\Omega}$ calibration load that is assigned zero delay by the manufacturer. The estimated delay is $38.8$ ps with a $1\\sigma$ uncertainty of $2.1$ ps for this particular load. This result is verified through measurements of a terminated airline. The measurements agree better with theoretical models of the airline when the reference plane is calibrated using th...

  8. Collaborative study for the calibration of a replacement International Standard for Tetanus Toxoid Adsorbed.

    Science.gov (United States)

    Tierney, Rob; Stickings, Paul; Hockley, Jason; Rigsby, Peter; Iwaki, Masaaki; Sesardic, Dorothea

    2011-11-01

    We present the results of a collaborative study for the establishment of a replacement International Standard (IS) for Tetanus Toxoid Adsorbed. Two candidate preparations were included in the study, one of which was established as the 4th IS for Tetanus Toxoid Adsorbed at the WHO Expert Committee on Biological Standardization meeting in October 2010. This preparation was found to have a unitage of 490 IU/ampoule, based on calibration in guinea pig challenge assays. Results from mouse challenge assays suggest that the relative performance of two candidate preparations may differ significantly between guinea pigs and mice. The authors note that the number of laboratories that performed guinea pig challenge assays, which are used to calibrate and assign IU, is much lower than in previous collaborative studies and this may have implications for calibration of replacement standards in the future. The issue of assigning separate units to the IS for guinea pig and mouse assays is discussed. The study also assessed performance of the replacement standard in serological assays which are used as alternative procedures to challenge assays for tetanus potency testing. Results suggest that the replacement standard is suitable for use as the reference vaccine in serological assays.

  9. A practical implementation of microphone free-field comparison calibration according to the standard IEC 61094-8

    DEFF Research Database (Denmark)

    Barrera Figueroa, Salvador; Torras-Rosell, Antoni; Rasmussen, Knud;

    2012-01-01

    An international standard concerned with the calibration of microphones in a free field by comparison has recently been published. The standard contemplates two main calibration methodologies for determining the sensitivity of a microphone under test when compared against a reference microphone. ...

  10. Standard test method for calibration of surface/stress measuring devices

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    1997-01-01

    Return to Contents page 1.1 This test method covers calibration or verification of calibration, or both, of surface-stress measuring devices used to measure stress in annealed and heat-strengthened or tempered glass using polariscopic or refractometry based principles. 1.2 This test method is nondestructive. 1.3 This test method uses transmitted light, and therefore, is applicable to light-transmitting glasses. 1.4 This test method is not applicable to chemically tempered glass. 1.5 Using the procedure described, surface stresses can be measured only on the “tin” side of float glass. 1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

  11. QconCAT standard for calibration of ion mobility-mass spectrometry systems.

    Science.gov (United States)

    Chawner, Ross; McCullough, Bryan; Giles, Kevin; Barran, Perdita E; Gaskell, Simon J; Eyers, Claire E

    2012-11-01

    Ion mobility-mass spectrometry (IM-MS) is a useful technique for determining information about analyte ion conformation in addition to mass/charge ratio. The physical principles that govern the mobility of an ion through a gas in the presence of a uniform electric field are well understood, enabling rotationally averaged collision cross sections (Ω) to be directly calculated from measured drift times under well-defined experimental conditions. However, such "first principle" calculations are not straightforward for Traveling Wave (T-Wave) mobility separations due to the range of factors that influence ion motion through the mobility cell. If collision cross section information is required from T-Wave mobility separations, then calibration of the instruments using known standards is essential for each set of experimental conditions. To facilitate such calibration, we have designed and generated an artificial protein based on the QconCAT technology, QCAL-IM, which upon proteolysis can be used as a universal ion mobility calibration standard. This single unique standard enables empirical calculation of peptide ion collision cross sections from the drift time on a T-Wave mobility instrument.

  12. Traceability of the ionization chambers calibrated at Kyushu regional center (I) for standard doses in medicine

    Energy Technology Data Exchange (ETDEWEB)

    Araki, Fujio; Noda, Hiroji; Higashida, Yoshiharu; Matsumoto, Masanori [Kumamoto Univ. (Japan). Coll. of Medical Science

    1997-12-01

    We investigated the traceability of the ionization chambers calibrated by the standard dosimeter with Kyushu regional center (I) for standard doses in medicine. Accuracy management of the ionization chambers of Kyushu regional center (I) has been maintained within 0.5% by intercomparison with that of the National Institute of Radiological Sciences (NIRS). The number of hospitals and the number of intercomparisons were 17 and 33 on an average per year, respectively. Accuracy of the ionization chambers of each hospital in recent years was less than 0.5% to the hospitals of 65-80% and less than 1.0% to those of 80-90%. (author)

  13. Dental Hygiene Faculty Calibration Using Two Accepted Standards for Calculus Detection: A Pilot Study.

    Science.gov (United States)

    Santiago, Lisa J; Freudenthal, Jacqueline J; Peterson, Teri; Bowen, Denise M

    2016-08-01

    Faculty calibration studies for calculus detection use two different standards for examiner evaluation, yet the only therapeutic modality that can be used for nonsurgical periodontal treatment is scaling/root debridement or planing. In this study, a pretest-posttest design was used to assess the feasibility of faculty calibration for calculus detection using two accepted standards: that established by the Central Regional Dental Testing Service, Inc. (CRDTS; readily detectible calculus) and the gold standard for scaling/root debridement (root roughness). Four clinical dental hygiene faculty members out of five possible participants at Halifax Community College agreed to participate. The participants explored calculus on the 16 assigned teeth (64 surfaces) of four patients. Calculus detection scores were calculated before and after training. Kappa averages using CRDTS criteria were 0.561 at pretest and 0.631 at posttest. Kappa scores using the scaling/root debridement or planing standard were 0.152 at pretest and 0.271 at posttest. The scores indicated improvement from moderate (Kappa=0.41-0.60) to substantial agreement (Kappa=0.61-0.80) following training using the CRDTS standard. Although this result differed qualitatively and Kappas were significantly different from 0, the differences for pre- to post-Kappas for patient-rater dyads using CRDTS were not statistically significant (p=0.778). There was no difference (p=0.913) in Kappa scores pre- to post-training using the scaling/root debridement standard. Despite the small number of participants in this study, the results indicated that training to improve interrater reliability to substantial agreement was feasible using the CRDTS standard but not using the gold standard. The difference may have been due to greater difficulty in attaining agreement regarding root roughness. Future studies should include multiple training sessions with patients using the same standard for scaling/root debridement used for

  14. Source geometry factors for HDR ¹⁹²Ir brachytherapy secondary standard well-type ionization chamber calibrations.

    Science.gov (United States)

    Shipley, D R; Sander, T; Nutbrown, R F

    2015-03-21

    Well-type ionization chambers are used for measuring the source strength of radioactive brachytherapy sources before clinical use. Initially, the well chambers are calibrated against a suitable national standard. For high dose rate (HDR) (192)Ir, this calibration is usually a two-step process. Firstly, the calibration source is traceably calibrated against an air kerma primary standard in terms of either reference air kerma rate or air kerma strength. The calibrated (192)Ir source is then used to calibrate the secondary standard well-type ionization chamber. Calibration laboratories are usually only equipped with one type of HDR (192)Ir source. If the clinical source type is different from that used for the calibration of the well chamber at the standards laboratory, a source geometry factor, k(sg), is required to correct the calibration coefficient for any change of the well chamber response due to geometric differences between the sources. In this work we present source geometry factors for six different HDR (192)Ir brachytherapy sources which have been determined using Monte Carlo techniques for a specific ionization chamber, the Standard Imaging HDR 1000 Plus well chamber with a type 70010 HDR iridium source holder. The calculated correction factors were normalized to the old and new type of calibration source used at the National Physical Laboratory. With the old Nucletron microSelectron-v1 (classic) HDR (192)Ir calibration source, ksg was found to be in the range 0.983 to 0.999 and with the new Isodose Control HDR (192)Ir Flexisource k(sg) was found to be in the range 0.987 to 1.004 with a relative uncertainty of 0.4% (k = 2). Source geometry factors for different combinations of calibration sources, clinical sources, well chambers and associated source holders, can be calculated with the formalism discussed in this paper.

  15. Phase calibration of sonar systems using standard targets and dual-frequency transmission pulses.

    Science.gov (United States)

    Islas-Cital, Alan; Atkins, Philip R; Foo, Kae Y; Picó, Ruben

    2011-10-01

    The phase angle component of the complex frequency response of a sonar system operating near transducer resonance is usually distorted. Interpretation and classification of the received sonar signal benefits from the preservation of waveform fidelity over the full bandwidth. A calibration process that measures the phase response in addition to the amplitude response is thus required. This paper describes an extension to the standard-target calibration method to include phase angle, without affecting the experimental apparatus, by using dual-frequency transmission pulses and frequency-domain data processing. This approach reduces the impact of unknown range and sound speed parameters upon phase calibration accuracy, as target phase is determined from the relationship of the two frequency components instead of relying on a local phase reference. Tungsten carbide spheres of various sizes were used to simultaneously calibrate the amplitude and phase response of an active sonar system in a laboratory tank. Experimental measurements of target phase spectra are in good agreement with values predicted from a theoretical model based upon full-wave analysis, over an operating frequency of 50-125 kHz.

  16. A standard stellar library for evolutionary synthesis, pt. III. Metallicity calibration

    CERN Document Server

    Westera, P; Buser, R; Cuisinier, F; Bruzual, A G

    2002-01-01

    We extend the colour calibration of the widely used BaSeL standard stellar library (Lejeune, Cuisinier, & Buser 1997, 1998) to non-solar metallicities, down to [Fe/H] ~ -2.0 dex. Surprisingly, we find that at the present epoch it is virtually impossible to establish a unique calibration of UBVRIJHKL colours in terms of stellar metallicity [Fe/H] which is consistent simultaneously with both colour-temperature relations and colour-absolute magnitude diagrams (CMDs) based on observed globular cluster photometry data and on published, currently popular standard stellar evolutionary tracks and isochrones. The problem appears to be related to the long-standing incompleteness in our understanding of convection in late-type stellar evolution, but is also due to a serious lack of relevant observational calibration data that would help resolve, or at least further significant progress towards resolving this issue. In view of the most important applications of the BaSeL library, we here propose two different metalli...

  17. Precise and accurate measurements of strong-field photoionisation and a transferrable laser intensity calibration standard

    CERN Document Server

    Wallace, W C; Khurmi, C; U., Satya Sainadh; Calvert, J E; Laban, D E; Pullen, M G; Bartschat, K; Grum-Grzhimailo, A N; Wells, D; Quiney, H M; Tong, X M; Litvinyuk, I V; Sang, R T; Kielpinski, D

    2016-01-01

    Ionization of atoms and molecules in strong laser fields is a fundamental process in many fields of research, especially in the emerging field of attosecond science. So far, demonstrably accurate data have only been acquired for atomic hydrogen (H), a species that is accessible to few investigators. Here we present measurements of the ionization yield for argon, krypton, and xenon with percentlevel accuracy, calibrated using H, in a laser regime widely used in attosecond science. We derive a transferrable calibration standard for laser peak intensity, accurate to 1.3%, that is based on a simple reference curve. In addition, our measurements provide a much-needed benchmark for testing models of ionisation in noble-gas atoms, such as the widely employed single-active electron approximation.

  18. Re-calibration of the NIST SRM 2059 master standard using traceable atomic force microscope metrology

    Science.gov (United States)

    Dixson, Ronald; Potzick, James; Orji, Ndubuisi G.

    2008-10-01

    The current photomask linewidth Standard Reference Material (SRM) supplied by the National Institute of Standards and Technology (NIST), SRM 2059, is the fifth generation of such standards for mask metrology. An in house optical microscope tool developed at NIST, called the NIST ultra-violet (UV) microscope, was used in transmission mode to calibrate the SRM 2059 photomasks. Due to the limitations of available optical models for determining the edge response in the UV microscope, the tool was used in a comparator mode. One of the masks was selected as a master standard - and the features on this mask were calibrated using traceable critical dimension atomic force microscope (CD-AFM) dimensional metrology. The optical measurements were then used to determine the relative offsets between the widths on the master standard and individual masks for sale to customers. At the time of these measurements, however, the uncertainties in the CD-AFM reference metrology on the master standard were larger than can now be achieved because the NIST single crystal critical dimension reference material (SCCDRM) project had not been completed. Using our CD-AFM at NIST, we have performed new measurements on the SRM 2059 master standard. The new AFM results are in agreement with the prior measurements and have expanded uncertainties approximately one fourth of those of the earlier results for sub-micrometer features. When the optical comparator data for customers masks are reanalyzed using these new AFM results, we expect to reduce the combined reported uncertainties for the linewidths on the actual SRMs by at least 40 % for the nominal 0.25 μm features.

  19. MO-D-BRD-04: NIST Air-Kerma Standard for Electronic Brachytherapy Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Mitch, M. [Nat’l Institute of Standards & Technology (United States)

    2015-06-15

    Electronic brachytherapy (eBT) has seen an insurgence of manufacturers entering the US market for use in radiation therapy. In addition to the established interstitial, intraluminary, and intracavitary applications of eBT, many centers are now using eBT to treat skin lesions. It is important for medical physicists working with electronic brachytherapy sources to understand the basic physics principles of the sources themselves as well as the variety of applications for which they are being used. The calibration of the sources is different from vendor to vendor and the traceability of calibrations has evolved as new sources came to market. In 2014, a new air-kerma based standard was introduced by the National Institute of Standards and Technology (NIST) to measure the output of an eBT source. Eventually commercial treatment planning systems should accommodate this new standard and provide NIST traceability to the end user. The calibration and commissioning of an eBT system is unique to its application and typically entails a list of procedural recommendations by the manufacturer. Commissioning measurements are performed using a variety of methods, some of which are modifications of existing AAPM Task Group protocols. A medical physicist should be familiar with the different AAPM Task Group recommendations for applicability to eBT and how to properly adapt them to their needs. In addition to the physical characteristics of an eBT source, the photon energy is substantially lower than from HDR Ir-192 sources. Consequently, tissue-specific dosimetry and radiobiological considerations are necessary when comparing these brachytherapy modalities and when making clinical decisions as a radiation therapy team. In this session, the physical characteristics and calibration methodologies of eBt sources will be presented as well as radiobiology considerations and other important clinical considerations. Learning Objectives: To understand the basic principles of electronic

  20. Distribution of radionuclides in an iron calibration standard for a free release measurement facility.

    Science.gov (United States)

    Hult, Mikael; Stroh, Heiko; Marissens, Gerd; Tzika, Faidra; Lutter, Guillaume; Šurán, Jiri; Kovar, Petr; Skala, Lukas; Sud, Jaromír

    2016-03-01

    A Europallet-sized calibration standard composed of 12 grey cast iron tubes contaminated with (60)Co and (110m)Ag with a mass of 246kg was developed. As the tubes were produced through centrifugal casting it was of particular concern to study the distribution of radionuclides in the radial direction of the tubes. This was done by removing 72 small samples (swarf) of ~0.3g each on both the inside and outside of the tubes. All of the samples were measured in the underground laboratory HADES.

  1. Electromagnetic coupling into two standard calibration shields on the Sandia cable tester

    Energy Technology Data Exchange (ETDEWEB)

    Warne, Larry Kevin; Basilio, Lorena I.; Langston, William L.; Chen, Kenneth C.; Hudson, Howard Gerald; Morris, M. E.; Stronach, S.L.; Johnson, W. A.; Derr, W.

    2014-02-01

    This report presents analytic transmission line models for calculating the shielding effectiveness of two common calibration standard cables. The two cables have different canonical aperture types, which produce the same low frequency coupling but different responses at resonance. The dominant damping mechanism is produced by the current probe loads at the ends of the cables, which are characterized through adaptor measurements. The model predictions for the cables are compared with experimental measurements and good agreement between the results is demonstrated. This setup constitutes a nice repeatable geometry that nevertheless exhibits some of the challenges involved in modeling non-radio frequency geometries.

  2. A practical implementation of microphone free-field comparison calibration according to the standard IEC 61094-8

    DEFF Research Database (Denmark)

    Barrera Figueroa, Salvador; Torras Rosell, Antoni; Rasmussen, Knud;

    2012-01-01

    An international standard concerned with the calibration of microphones in a free field by comparison has recently been published. The standard contemplates two main calibration methodologies for determining the sensitivity of a microphone under test when compared against a reference microphone....... A third method, consisting of a combination of the sequential and simultaneous methodologies, has also been investigated. Though the application of time selective techniques is not discussed, the experimental results indicate the immunity to unwanted reflections in the sequential and combined approaches...

  3. Production and characterization of a nitrogen-implanted Fe standard to calibrate PIGE measurements

    Science.gov (United States)

    Rodrigues, C. L.; Silva, T. F.; Added, N.; Santos, H. C.; Tabacniks, M. H.

    2014-11-01

    Three calibration standard was produced by ion implantation of nitrogen in samples of Armco iron (99.7% iron). The samples was irradiated with nitrogen ion beams at several different energies (between 4 keV and 40 keV), and the ion doses were adjusted to obtain an uniform depth profile, using simulations with SRIM code. Two standards, one thick and other a foil (1.62mg/cm2), was irradiated at same time with total nominal dose of 6.6×10-16 atoms/cm2 distributed in a region of 100 nm in depth, with an average concentration of 9.0% nitrogen in iron. The third sample uses the same profile, but with a small dose, 1.1×10-16 atoms/cm2 and average concentration of 1.5% nitrogen. The characterization of the implanted samples was done using RBS and NRA techniques to quantification of nitrogen.

  4. Compensation for the Effects of Ambient Conditions on the Calibration of Multi-Capillary Pressure Drop Standards

    Directory of Open Access Journals (Sweden)

    Colard S

    2014-12-01

    Full Text Available Cigarette draw resistance and filter pressure drop (PD are both major physical parameters for the tobacco industry. Therefore these parameters must be measured reliably. For these measurements, specific equipment calibrated with PD transfer standards is used. Each transfer standard must have a known and stable PD value, such standards usually being composed of several capillary tubes associated in parallel. However, PD values are modified by ambient conditions during calibration of such standards, i.e. by temperature and relative humidity (RH of air, and atmospheric pressure. In order to reduce the influence of these ambient factors, a simplified model was developed for compensating the effects of ambient conditions on the calibration of multi-capillary PD standards.

  5. Airborne hygrometer calibration inter-comparison against a metrological water vapour standard

    Science.gov (United States)

    Smorgon, Denis; Boese, Norbert; Ebert, Volker

    2014-05-01

    Water vapour is the most important atmospheric greenhouse gas, which causes a major feedback to warming and other changes in the climate system. Knowledge of the distribution of water vapour and its climate induced changes is especially important in the upper troposphere and lower stratosphere (UT/LS) where vapour plays a critical role in atmospheric radiative balance, cirrus cloud formation, and photochemistry. But, our understanding of water in the UT/LS is limited by significant uncertainties in current UT/LS water measurements. One of the most comprehensive inter-comparison campaigns for airborne hygrometers, termed AQUAVIT (AV1) [1], took place in 2007 at the AIDA chamber at the Karlsruhe Institute of Technology (KIT) in Germany. AV1 was a well-defined, referred, blind inter-comparison of 22 airborne field instruments from 17 international research groups. One major metrological deficit of AV1, however, was, that no traceable reference instrument participated in the inter-comparison experiments and that the calibration procedures of the participating instruments were not monitored or interrogated. Consequently a follow-up inter-comparison was organized in April 2013, which for the first time also provides a traceable link to the international humidity scale. This AQUAVIT2 (AV2) campaign (details see: http://www.imk-aaf.kit.edu/aquavit/index.php/Main_Page) was again located at KIT/AIDA and organised by an international organizing committee including KIT, PTB, FZJ and others. Generally AV2 is divided in two parallel comparisons: 1) AV2-A uses the AIDA chamber for a simultaneous comparison of all instruments (incl. sampling and in-situ instruments) over a broad range of conditions characteristic for the UT/LS; 2) AV2-B, about which this paper is reporting, is a sequential comparison of selected hygrometers and (when possible) their reference calibration infrastructures by means of a chilled mirror hygrometer traced back to the primary National humidity standard

  6. Binary Pseudo-random Grating Standard for Calibration of Surface Profilometers

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy; Yashchuk, Valeriy V.; McKinney, Wayne R.; Takacs, Peter Z.

    2008-01-16

    We suggest and describe the use of a binary pseudo-random (BPR) grating as a standard test surface for measurement of the modulation transfer function (MTF) of interferometric microscopes. Knowledge of the MTF of a microscope is absolutely necessary to convert the measured height distribution of a surface undergoing metrology into an accurate power spectral density (PSD) distribution. For an'ideal' microscope with an MTF function independent of spatial frequency out to the Nyquist frequency of the detector array with zero response at higher spatial frequencies, a BPR grating would produce a flat 1D PSD spectrum, independent of spatial frequency. For a'real' instrument, the MTF is found as the square root of the ratio of the PSD spectrum measured with the BPR grating to the'ideal,' spatial frequency independent, PSD spectrum. We present the results from a measurement of the MTF of MicromapTM-570 interferometric microscope demonstrating a high efficiency for the calibration method.

  7. Calibration and standards beamline 6.3.2 at the ALS

    Energy Technology Data Exchange (ETDEWEB)

    Underwood, J.H.; Gullikson, E.M.; Koike, M. [Ernest Orlando Lawrence Berkeley National Lab., CA (United States)] [and others

    1997-04-01

    More sophisticated optics for the x-ray, soft x-ray and far ultraviolet spectral regions being developed for synchrotron radiation research and many other applications, require accurate calibration and standards facilities for measuring reflectivity of mirrors and multilayer coatings, transmission of thin films, bandpass of multilayers, efficiency of gratings or detectors, etc. For this purpose beamline 6.3.2 was built at the ALS. Its energy coverage, versatility, simplicity and convenience also make it useful for a wide range of other experiments. The paper describes the components of this beamline, consisting of: a four jaw aperture; a horizontal focusing mirror; a monochromator; exit slit; vertical focusing mirror; mechanical and vacuum system; reflectometer; filter wheels; and data acquisition system.

  8. Production and characterization of a nitrogen-implanted Fe standard to calibrate PIGE measurements

    Energy Technology Data Exchange (ETDEWEB)

    Rodrigues, C. L.; Silva, T. F.; Added, N.; Santos, H. C.; Tabacniks, M. H. [Institute of Physics - University of São Paulo (Brazil)

    2014-11-11

    Three calibration standard was produced by ion implantation of nitrogen in samples of Armco iron (99.7% iron). The samples was irradiated with nitrogen ion beams at several different energies (between 4 keV and 40 keV), and the ion doses were adjusted to obtain an uniform depth profile, using simulations with SRIM code. Two standards, one thick and other a foil (1.62mg/cm{sup 2}), was irradiated at same time with total nominal dose of 6.6×10{sup −16} atoms/cm{sup 2} distributed in a region of 100 nm in depth, with an average concentration of 9.0% nitrogen in iron. The third sample uses the same profile, but with a small dose, 1.1×10{sup −16} atoms/cm{sup 2} and average concentration of 1.5% nitrogen. The characterization of the implanted samples was done using RBS and NRA techniques to quantification of nitrogen.

  9. Calibration standard of body tissue with magnetic nanocomposites for MRI and X-ray imaging

    Science.gov (United States)

    Rahn, Helene; Woodward, Robert; House, Michael; Engineer, Diana; Feindel, Kirk; Dutz, Silvio; Odenbach, Stefan; StPierre, Tim

    2016-05-01

    We present a first study of a long-term phantom for Magnetic Resonance Imaging (MRI) and X-ray imaging of biological tissues with magnetic nanocomposites (MNC) suitable for 3-dimensional and quantitative imaging of tissues after, e.g. magnetically assisted cancer treatments. We performed a cross-calibration of X-ray microcomputed tomography (XμCT) and MRI with a joint calibration standard for both imaging techniques. For this, we have designed a phantom for MRI and X-ray computed tomography which represents biological tissue enriched with MNC. The developed phantoms consist of an elastomer with different concentrations of multi-core MNC. The matrix material is a synthetic thermoplastic gel, PermaGel (PG). The developed phantoms have been analyzed with Nuclear Magnetic Resonance (NMR) Relaxometry (Bruker minispec mq 60) at 1.4 T to obtain R2 transverse relaxation rates, with SQUID (Superconducting QUantum Interference Device) magnetometry and Inductively Coupled Plasma Mass Spectrometry (ICP-MS) to verify the magnetite concentration, and with XμCT and 9.4 T MRI to visualize the phantoms 3-dimensionally and also to obtain T2 relaxation times. A specification of a sensitivity range is determined for standard imaging techniques X-ray computed tomography (XCT) and MRI as well as with NMR. These novel phantoms show a long-term stability over several months up to years. It was possible to suspend a particular MNC within the PG reaching a concentration range from 0 mg/ml to 6.914 mg/ml. The R2 relaxation rates from 1.4 T NMR-relaxometry show a clear connection (R2=0.994) with MNC concentrations between 0 mg/ml and 4.5 mg/ml. The MRI experiments have shown a linear correlation of R2 relaxation and MNC concentrations as well but in a range between MNC concentrations of 0 mg/ml and 1.435 mg/ml. It could be shown that XμCT displays best moderate and high MNC concentrations. The sensitivity range for this particular XμCT apparatus yields from 0.569 mg/ml to 6.914 mg/ml. The

  10. Development and validation of a portable gas phase standard generation and calibration system for volatile organic compounds

    Directory of Open Access Journals (Sweden)

    P. Veres

    2010-01-01

    Full Text Available We report on the development of an accurate, portable, dynamic calibration system for volatile organic compounds (VOCs. The Mobile Organic Carbon Calibration System (MOCCS combines the production of gas-phase VOC standards using permeation or diffusion sources with quantitative total organic carbon (TOC conversion on a palladium surface to CO2 in the presence of oxygen, and the subsequent CO2 measurement. MOCCS was validated using three different comparisons: (1 TOC of high accuracy methane standards compared well to expected concentrations (3% relative error, (2 a gas-phase benzene standard was generated using a permeation source and measured by TOC and gas chromatography mass spectrometry (GC-MS with excellent agreement (<4% relative difference, and (3 total carbon measurement of 4 known gas phase mixtures were performed and compared to a calculated carbon content to agreement within the stated uncertainties of the standards. Measurements from laboratory biomass burning experiments of formic acid by negative-ion proton-transfer chemical-ionization mass spectrometry (NI-PT-CIMS and formaldehyde by proton transfer reaction-mass spectrometry (PTR-MS, both calibrated using MOCCS, were compared to open path Fourier transform infrared spectroscopy (OP-FTIR to validate the MOCCS calibration and were found to compare well (R2 of 0.91 and 0.99 respectively.

  11. Development and validation of a portable gas phase standard generation and calibration system for volatile organic compounds

    Directory of Open Access Journals (Sweden)

    P. Veres

    2010-06-01

    Full Text Available We report on the development of an accurate, portable, dynamic calibration system for volatile organic compounds (VOCs. The Mobile Organic Carbon Calibration System (MOCCS combines the production of gas-phase VOC standards using permeation or diffusion sources with quantitative total organic carbon (TOC conversion on a palladium surface to CO2 in the presence of oxygen, and the subsequent CO2 measurement. MOCCS was validated using three different comparisons: (1 TOC of high accuracy methane standards compared well to expected concentrations (3% relative error, (2 a gas-phase benzene standard was generated using a permeation source and measured by TOC and gas chromatography mass spectrometry (GC-MS with excellent agreement (<4% relative difference, and (3 total carbon measurement of 4 known gas phase mixtures were performed and compared to a calculated carbon content to agreement within the stated uncertainties of the standards. Measurements from laboratory biomass burning experiments of formic acid by negative-ion proton-transfer chemical-ionization mass spectrometry (NI-PT-CIMS and formaldehyde by proton transfer reaction-mass spectrometry (PTR-MS, both calibrated using MOCCS, were compared to open path Fourier transform infrared spectroscopy (OP-FTIR to validate the MOCCS calibration and were found to compare well (R2 of 0.91 and 0.99, respectively.

  12. Developing Calibration Weights and Standard-Error Estimates for a Survey of Drug-Related Emergency-Department Visits

    Directory of Open Access Journals (Sweden)

    Kott Phillip S.

    2014-09-01

    Full Text Available This article describes a two-step calibration-weighting scheme for a stratified simple random sample of hospital emergency departments. The first step adjusts for unit nonresponse. The second increases the statistical efficiency of most estimators of interest. Both use a measure of emergency-department size and other useful auxiliary variables contained in the sampling frame. Although many survey variables are roughly a linear function of the measure of size, response is better modeled as a function of the log of that measure. Consequently the log of size is a calibration variable in the nonresponse-adjustment step, while the measure of size itself is a calibration variable in the second calibration step. Nonlinear calibration procedures are employed in both steps. We show with 2010 DAWN data that estimating variances as if a one-step calibration weighting routine had been used when there were in fact two steps can, after appropriately adjusting the finite-population correct in some sense, produce standard-error estimates that tend to be slightly conservative.

  13. Absolute Flux Calibration of the IRAC Instrument on the Spitzer Space Telescope using Hubble Space Telescope Flux Standards

    CERN Document Server

    Bohlin, R C; Rieke, G H; Ardila, D; Carey, S; Deustua, S; Engelbracht, C; Ferguson, H C; Flanagan, K; Kalirai, J; Meixner, M; Noriega-Crespo, A; Su, K Y L; Tremblay, P -E

    2011-01-01

    The absolute flux calibration of the James Webb Space Telescope will be based on a set of stars observed by the Hubble and Spitzer Space Telescopes. In order to cross-calibrate the two facilities, several A, G, and white dwarf (WD) stars are observed with both Spitzer and Hubble and are the prototypes for a set of JWST calibration standards. The flux calibration constants for the four Spitzer IRAC bands 1-4 are derived from these stars and are 2.3, 1.9, 2.0, and 0.5% lower than the official cold-mission IRAC calibration of Reach et al. (2005), i.e. in agreement within their estimated errors of ~2%. The causes of these differences lie primarily in the IRAC data reduction and secondarily in the SEDs of our standard stars. The independent IRAC 8 micron band-4 fluxes of Rieke et al. (2008) are about 1.5 +/- 2% higher than those of Reach et al. and are also in agreement with our 8 micron result.

  14. Development of transfer standard devices for ensuring the accurate calibration of ultrasonic physical therapy machines in clinical use

    Energy Technology Data Exchange (ETDEWEB)

    Hekkenberg, R T [TNO Prevention and Health, Zernikedreef 9, 2333 CK Leiden (Netherlands); Richards, A [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Beissner, K [Physikalisch-Technische Bundesanstalt, PTB, Bundesallee 100, D-38116 Braunschweig (Germany); Zeqiri, B [National Physical Laboratory, NPL, Queens Road, Teddington, TW11 0LW (United Kingdom); Prout, G [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Cantrall, Ch [National Measurement Laboratory, CSIRO, Bradfield Rd, West Lindfield 2070, Sydney (Australia); Bezemer, R A [TNO Prevention and Health, Zernikedreef 9, 2333 CK Leiden (Netherlands); Koch, Ch [Physikalisch-Technische Bundesanstalt, PTB, Bundesallee 100, D-38116 Braunschweig, (Germany); Hodnett, M [National Physical Laboratory, NPL, Queens Road, Teddington, TW11 0LW (United Kingdom)

    2004-01-01

    Physical therapy ultrasound is widely applied to patients. However, many devices do not comply with the relevant standard stating that the actual power output shall be within {+-}20% of the device indication. Extreme cases have been reported: from delivering effectively no ultrasound or operating at maximum power at all powers indicated. This can potentially lead to patient injury as well as mistreatment. The present European (EC) project is an ongoing attempt to improve the quality of the treatment of patients being treated with ultrasonic physical-therapy. A Portable ultrasound Power Standard (PPS) is being developed and accurately calibrated. The PPS includes: Ultrasound transducers (including one exhibiting an unusual output) and a driver for the ultrasound transducers that has calibration and proficiency test functions. Also included with the PPS is a Cavitation Detector to determine the onset of cavitation occurring within the propagation medium. The PPS will be suitable for conducting in-the-field accreditation (proficiency testing and calibration). In order to be accredited it will be important to be able to show traceability of the calibration, the calibration process and qualification of testing staff. The clinical user will benefit from traceability because treatments will be performed more reliably.

  15. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration unce...

  16. A new standard cylindrical graphite-walled ionization chamber for dosimetry in 60Co beams at calibration laboratories

    Science.gov (United States)

    Neves, Lucio P.; Perini, Ana P.; Caldas, Linda V. E.

    2014-11-01

    60Co sources are used mostly at dosimetry laboratories for calibration of ionization chambers utilized for radiotherapy dosimetry, mainly in those laboratories where there is no linear accelerator available. In this work, a new cylindrical ionization chamber was developed and characterized to be used as a reference dosimeter at the Calibration Laboratory of the IPEN. The characterization tests were performed according to the IEC 60731 standard, and all tests presented results within its recommended limits. Furthermore, the correction factors for the wall, stem, central collecting electrode, nonaxial uniformity and the mass-energy absorption coefficient were determined using the EGSnrc Monte Carlo code. The air kerma rate determined with this new dosimeter was compared to the one obtained with the IPEN standard, presenting a difference of 1.5%. Therefore, the new ionization chamber prototype developed and characterized in this work presents potential use as a primary standard dosimeter at radiation metrology laboratories.

  17. Flux Calibration of the BATC Survey and Its Application for Checking the Consistency of the Oke-Gunn Standards

    Institute of Scientific and Technical Information of China (English)

    2001-01-01

    Four Oke-Gunn (OG) standards, HD 19445, HD 84937, BD+26 2606 and BD+17 4708 are used as standard stars for flux calibration in the BATC project. They are also widely used in the visual wavelength region in many other photometric projects. Over the years we have observed on 58 good photometric nights, and the data obtained are used for flux calibration. Normally two or three OG standards are observed in every photometric night. The data are used for getting the atmo-spheric extinction coefficients and instrumental magnitude zero point. We also use these data to make inter-comparisons among the magnitudes of these standard stars.As a result, we found the magnitudes of HD 19445, HD 84937 and BD+17 4708 to agree well with those estimated in previous work to within 0.03 magnitude. How-ever, BD+26 2606 shows a larger deviation especially at short wavebands. Possible reasons are analyzed and the revised magnitudes are obtained for these standards.It is shown that the quality of flux calibration of the BATC fields is significantly improved by applying the new magnitudes.

  18. Spectral Irradiance Calibration in the Infrared. 11; Comparison of (alpha) Bootis and 1 Ceres with a Laboratory Standard

    Science.gov (United States)

    Witteborn, Fred C.; Cohen, Martin; Bregman, Jesse D.; Wooden, Diane H.; Heere, Karen; Shirley, Eric L.

    1999-01-01

    Infrared spectra of two celestial objects frequently used as flux standards are calibrated against an absolute laboratory flux standard at a spectral resolving power of 100 to 200. The spectrum of the KI.5 III star alpha Boo is measured from 3 to 30 microns, and that of the C-type asteroid 1 Ceres from 5 to 30 microns. While these "standard" spectra do not have the apparent precision of those based on calculated models, they do not require the assumptions involved in theoretical models of stars and asteroids. Specifically, they provide a model-independent means of calibrating celestial flux in the spectral range from 12 to 30 microns, where accurate absolute photometry is not available. The agreement found between the spectral shapes of alpha Boo and Ceres based on laboratory standards and those based on observed ratios to alpha CMa (Sirius) and alpha Lyr (Vega), flux-calibrated by theoretical modeling of these hot stars, strengthens our confidence in the applicability of the stellar models as primary irradiance standards.

  19. Spectral Irradiance Calibration in the Infrared 11: Comparison of (alpha) Boo and 1 Ceres with a Laboratory Standard

    Science.gov (United States)

    Witteborn, Fred C.; Cohen, Martin; Bregman, Jess D.; Wooden, Diane; Heere, Karen; Shirley, Eric L.

    1998-01-01

    Infrared spectra of two celestial objects frequently used as flux standards are calibrated against an absolute laboratory flux standard at a spectral resolving power of 100 to 200. The spectrum of the K1.5III star, alpha Boo, is measured from 3 microns to 30 microns and that of the C-type asteroid, 1 Ceres, from 5 microns to 30 microns. While these 'standard' spectra do not have the apparent precision of those based on calculated models, they do not require the assumptions involved in theoretical models of stars and asteroids. Specifically they provide a model-independent means of calibrating celestial flux in the spectral range from 12 microns to 30 microns where accurate absolute photometry is not available. The agreement found between the spectral shapes of alpha Boo and Ceres based on laboratory standards, and those based on observed ratios to alpha CMa (Sirius) and alpha Lyr (Vega), flux calibrated by theoretical modeling of these hot stars strengthens our confidence in the applicability of the stellar models as primary irradiance standards.

  20. The use of calibration standards and the correction for sample self-attenuation in gamma-ray nondestructive assay

    Energy Technology Data Exchange (ETDEWEB)

    Parker, J.L.

    1986-11-01

    The efficient use of appropriate calibration standards and the correction for the attenuation of the gamma rays within an assay sample by the sample itself are two important and closely related subjects in gamma-ray nondestructive assay. Much research relating to those subjects has been done in the Nuclear Safeguards Research and Development program at the Los Alamos National Laboratory since 1970. This report brings together most of the significant results of that research. Also discussed are the nature of appropriate calibration standards and the necessary conditions on the composition, size, and shape of the samples to allow accurate assays. Procedures for determining the correction for the sample self-attenuation are described at length including both general principles and several specific useful cases. The most useful concept is that knowing the linear attenuation coefficient of the sample (which can usually be determined) and the size and shape of the sample and its position relative to the detector permits the computation of the correction factor for the self-attenuation. A major objective of the report is to explain how the procedures for determining the self-attenuation correction factor can be applied so that calibration standards can be entirely appropriate without being particularly similar, either physically or chemically, to the items to be assayed. This permits minimization of the number of standards required to assay items with a wide range of size, shape, and chemical composition.

  1. Use of calibration standards and the correction for sample self-attenuation in gamma-ray nondestructive assay

    Energy Technology Data Exchange (ETDEWEB)

    Parker, J.L.

    1984-08-01

    The efficient use of appropriate calibration standards and the correction for the attenuation of the gamma rays within an assay sample by the sample itself are two important and closely related subjects in gamma-ray nondestructive assay. Much research relating to those subjects has been done in the Nuclear Safeguards Research and Development program at the Los Alamos National Laboratory since 1970. This report brings together most of the significant results of that research. Also discussed are the nature of appropriate calibration standards and the necessary conditions on the composition, size, and shape of the samples to allow accurate assays. Procedures for determining the correction for the sample self-attenuation are described at length including both general principles and several specific useful cases. The most useful concept is that knowing the linear attenuation coefficient of the sample (which can usually be determined) and the size and shape of the sample and its position relative to the detector permits the computation of the correction factor for the self-attenuation. A major objective of the report is to explain how the procedures for determining the self-attenuation correction factor can be applied so that calibration standards can be entirely appropriate without being particularly similar, either physically or chemically, to the items to be assayed. This permits minimization of the number of standards required to assay items with a wide range of size, shape, and chemical composition. 17 references, 18 figures, 2 tables.

  2. Free-field reciprocity calibration of laboratory standard (LS) microphones using a time selective technique

    DEFF Research Database (Denmark)

    Rasmussen, Knud; Barrera Figueroa, Salvador

    2006-01-01

    Although the basic principle of reciprocity calibration of microphones in a free field is simple, the practical problems are complicated due to the low signal-to-noise ratio and the influence of cross talk and reflections from the surroundings. The influence of uncorrelated noise can be reduced...

  3. EASYTRAC Project: Work package 6.4 Reversal technique to calibrate gear and thread standards

    DEFF Research Database (Denmark)

    Carmignato, Simone; De Chiffre, Leonardo; Larsen, Erik;

    This report was produced as a part of the project EASYTRAC, an EU project under the programme Competitive and Sustainable Growth: Contract No. G6RD-CT-2000-00188, coordinated by UNIMETRIK S.A. (Spain). The project is concerned with low uncertainty calibrations on coordinate measuring machines (CM...

  4. EDTA Standard Solution Calibration Factors Discussed in This Paper Gas Pipe Network in Service%EDTA标准溶液标定影响因素探讨

    Institute of Scientific and Technical Information of China (English)

    吴卫平

    2016-01-01

    本文分析EDTA标准溶液标定过程,以0.010.1mol/L EDTA标准溶液的标定过程影响因素进行分析探讨,以提高EDTA标准溶液标定的准确性。%In this paper, the EDTA standard solution calibration process to 0.01 mol/L EDTA standard solution, this paper analyzes the factors which affect the process of calibration, to improve the accuracy of the EDTA standard solution calibration.

  5. Loop system for creating jet fuel vapor standards used in the calibration of infrared spectrophotometers and gas chromatographs.

    Science.gov (United States)

    Reboulet, James; Cunningham, Robert; Gunasekar, Palur G; Chapman, Gail D; Stevens, Sean C

    2009-02-01

    A whole body inhalation study of mixed jet fuel vapor and its aerosol necessitated the development of a method for preparing vapor only standards from the neat fuel. Jet fuel is a complex mixture of components which partitions between aerosol and vapor when aspirated based on relative volatility of the individual compounds. A method was desired which could separate the vapor portion from the aerosol component to prepare standards for the calibration of infrared spectrophotometers and a head space gas chromatography system. A re-circulating loop system was developed which provided vapor only standards whose composition matched those seen in an exposure system. Comparisons of nominal concentrations in the exposure system to those determined by infrared spectrophotometry were in 92-95% agreement. Comparison of jet fuel vapor concentrations determined by infrared spectrophotometry compared to head space gas chromatography yielded a 93% overall agreement in trial runs. These levels of agreement show the loop system to be a viable method for creating jet fuel vapor standards for calibrating instruments.

  6. A standard stellar library for evolutionary synthesis; 1, Calibration of theoretical spectra

    CERN Document Server

    Lejeune, T; Buser, R

    1997-01-01

    A comprehensive hybrid library of synthetic stellar spectra based on three original grids of model atmosphere spectra by Kurucz (1995), Fluks et al. (1994), and Bessell et al. (1989, 1991) is presented. The combined library is intended for multiple-purpose synthetic photometry applications and is constructed in order (i) to cover the largest possible ranges in Teff, log g, and [M/H]), (ii) to provide flux spectra with useful resolution on the uniform grid of wavelengths, and (iii) to provide realistic synthetic broad-band colors for the largest possible parameter and wavelength ranges. For each value of the effective temperature and for each wavelength, we calculate the correction function that must be applied to a (theoretical) solar-abundance model flux spectrum which yields synthetic UBVRIJHKL colors matching the (empirical) color-temperature calibrations derived from observations. The calibration algorithm is designed to preserve the original differential grid properties implied by metallicity and/or lumi...

  7. Variations in performance of LCDs are still evident after DICOM gray-scale standard display calibration.

    LENUS (Irish Health Repository)

    Lowe, Joanna M

    2010-07-01

    Quality assurance in medical imaging is directly beneficial to image quality. Diagnostic images are frequently displayed on secondary-class displays that have minimal or no regular quality assurance programs, and treatment decisions are being made from these display types. The purpose of this study is to identify the impact of calibration on physical and psychophysical performance of liquid crystal displays (LCDs) and the extent of potential variance across various types of LCDs.

  8. Development and calibration of a standard for the protein content of granulocyte colony-stimulating factor products.

    Science.gov (United States)

    Gao, Kai; Rao, Chunming; Tao, Lei; Han, Chunmei; Shi, Xinchang; Wang, Lan; Fan, Wenhong; Yu, Lei; Wang, Junzhi

    2012-03-01

    This collaborative study characterizes a homogeneous standard for the protein content determination of granulocyte colony-stimulating factor (G-CSF) products with traceability of the measurement. The Kjeldahl method was used to determine the average protein content of G-CSF bulk as 2.505 mg/ml (95% C.I: 2.467-2.543 mg/ml, GCV 4.0%). Using G-CSF bulk as a traceability benchmark, the protein content of the final freeze-dried standard using reverse phase HPLC (RP-HPLC) was 215.4 μg protein per ampoule (95% C.I: 212.407-218.486 μg/ampoule, GCV 3.4%). A comparative study showed that there was no difference between using Filgrastim CRS (European Pharmacopeia G-CSF reference standard) and freeze-dried homogeneous standard when quantifying G-CSF protein content by RP-HPLC (P > 0.05). However, there were significant differences in the G-CSF protein content obtained using a serum albumin standard by Lowry assay and a G-CSF standard with RP-HPLC. Therefore, use of RP-HPLC with a freeze-dried homogeneous standard would eliminate the systematic errors introduced when using a serum albumin standard because of the differences in protein composition between the standard and the sample. It would also be helpful to use this method to compare the quality of G-CSF biosimilar products in situations where the protein content has been calibrated using various standards.

  9. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina

    2016-05-02

    This poster presents the development, implementation, and operation of the Broadband Outdoor Radiometer Calibrations (BORCAL) Longwave (LW) system at the Southern Great Plains Radiometric Calibration Facility for the calibration of pyrgeometers that provide traceability to the World Infrared Standard Group.

  10. (18)F primary standard at ENEA-INMRI by three absolute techniques and calibration of a well-type IG11 ionization chamber.

    Science.gov (United States)

    Capogni, Marco; Carconi, Pierluigi; De Felice, Pierino; Fazio, Aldo

    2016-03-01

    A new (18)F primary standardization carried out at ENEA-INMRI by three different absolute techniques, i.e. 4πγNaI(Tl)γ high-efficiency counting, TDCR and 4πβ(LS)-γ[NaI(Tl)] coincidence counting method, allowed the calibration of a fixed well-reentrant IG11 ionization chamber (IC), with an uncertainty lower than 1%, and to check the calibration factor of a portable well-type IC NPL-CRC model, previously calibrated. By the new standard the ENEA-INMRI was linked to the BIPM International Reference System (SIR) through the BIPM SIR Transfer Instrument (SIRTI).

  11. Development and evaluation of a set of group delay standards. [deep space tracking station calibration

    Science.gov (United States)

    Otoshi, T. Y.; Beatty, R. W.

    1976-01-01

    A set of cable assemblies serving as group delay standards having nominal delays of 15, 30, and 60 nsec are described. Various types of measurements were performed on the cable standards, including impedance, microwave phase shift, RF pulse burst delay, modulation pulsed delay, and envelope phase shift measurements. The results of these tests are given, and various sources of error are discussed, in particular, dispersion and internal reflections.

  12. A versatile and reproducible automatic injection system for liquid standard introduction: application to in-situ calibration

    Directory of Open Access Journals (Sweden)

    G. Isaacman

    2011-05-01

    Full Text Available The quantitation of trace organic compounds in ambient organic aerosol is difficult due to the chemical complexity of these mixtures, but is needed to provide insight into their sources and formation processes. Compound-level characterization of organic aerosols is typically performed through sample collection followed by gas or liquid chromatography. With these methods, introduction of liquid standards has long been used as an effective means of quantifying trace compounds, but automating this technique for use with in-situ instrumentation has not previously been achieved. Here we develop an automatic injection system (AutoInject for the introduction of liquids into a custom collection and analysis cell for improved quantitation in chromatographic measurements. The system consists of chilled reservoirs containing liquid standards from which a sample loop is loaded and then injected into the cell. The AutoInject is shown to have reproducibility over 106 injections with a relative standard deviation of 1.5 %, and have negligible injection-to-injection carryover. A 6-port selector allows injection of different liquid standards separately or simultaneously. Additionally, automatic injection of multiple sample loops is shown to generate a linear multi-point calibration curve. Tests conducted in this work focus on use with the Thermal desorption Aerosol Gas chromatograph (TAG, but the flexibility of the system allows it to be used for a variety of applications.

  13. Development, improvement and calibration of neutronic reaction rate measurements: elaboration of a base of standard techniques; Developpement, amelioration et calibration des mesures de taux de reaction neutroniques: elaboration d`une base de techniques standards

    Energy Technology Data Exchange (ETDEWEB)

    Hudelot, J.P

    1998-06-19

    In order to improve and to validate the neutronic calculation schemes, perfecting integral measurements of neutronic parameters is necessary. This thesis focuses on the conception, the improvement and the development of neutronic reaction rates measurements, and aims at building a base of standard techniques. Two subjects are discussed. The first one deals with direct measurements by fission chambers. A short presentation of the different usual techniques is given. Then, those last ones are applied through the example of doubling time measurements on the EOLE facility during the MISTRAL 1 experimental programme. Two calibration devices of fission chambers are developed: a thermal column located in the central part of the MINERVE facility, and a calibration cell using a pulsed high flux neutron generator and based on the discrimination of the energy of the neutrons with a time-of-flight method. This second device will soon allow to measure the mass of fission chambers with a precision of about 1 %. Finally, the necessity of those calibrations will be shown through spectral indices measurements in core MISTRAL 1 (UO{sub 2}) and MISTRAL 2 (MOX) of the EOLE facility. In each case, the associated calculation schemes, performed using the Monte Carlo MCNP code with the ENDF-BV library, will be validated. Concerning the second one, the goal is to develop a method for measuring the modified conversion ratio of {sup 238}U (defined as the ratio of {sup 238}U capture rate to total fission rate) by gamma-ray spectrometry of fuel rods. Within the framework of the MISTRAL 1 and MISTRAL 2 programmes, the measurement device, the experimental results and the spectrometer calibration are described. Furthermore, the MCNP calculations of neutron self-shielding and gamma self-absorption are validated. It is finally shown that measurement uncertainties are better than 1 %. The extension of this technique to future modified conversion ratio measurements for {sup 242}Pu (on MOX rods) and

  14. Calibration of sound calibrators: an overview

    Science.gov (United States)

    Milhomem, T. A. B.; Soares, Z. M. D.

    2016-07-01

    This paper presents an overview of calibration of sound calibrators. Initially, traditional calibration methods are presented. Following, the international standard IEC 60942 is discussed emphasizing parameters, target measurement uncertainty and criteria for conformance to the requirements of the standard. Last, Regional Metrology Organizations comparisons are summarized.

  15. Contributed Review: Absolute spectral radiance calibration of fiber-optic shock-temperature pyrometers using a coiled-coil irradiance standard lamp

    Energy Technology Data Exchange (ETDEWEB)

    Fat’yanov, O. V., E-mail: fatyan1@gps.caltech.edu; Asimow, P. D., E-mail: asimow@gps.caltech.edu [Division of Geological and Planetary Sciences 252-21, California Institute of Technology, Pasadena, California 91125 (United States)

    2015-10-15

    We describe an accurate and precise calibration procedure for multichannel optical pyrometers such as the 6-channel, 3-ns temporal resolution instrument used in the Caltech experimental geophysics laboratory. We begin with a review of calibration sources for shock temperatures in the 3000-30 000 K range. High-power, coiled tungsten halogen standards of spectral irradiance appear to be the only practical alternative to NIST-traceable tungsten ribbon lamps, which are no longer available with large enough calibrated area. However, non-uniform radiance complicates the use of such coiled lamps for reliable and reproducible calibration of pyrometers that employ imaging or relay optics. Careful analysis of documented methods of shock pyrometer calibration to coiled irradiance standard lamps shows that only one technique, not directly applicable in our case, is free of major radiometric errors. We provide a detailed description of the modified Caltech pyrometer instrument and a procedure for its absolute spectral radiance calibration, accurate to ±5%. We employ a designated central area of a 0.7× demagnified image of a coiled-coil tungsten halogen lamp filament, cross-calibrated against a NIST-traceable tungsten ribbon lamp. We give the results of the cross-calibration along with descriptions of the optical arrangement, data acquisition, and processing. We describe a procedure to characterize the difference between the static and dynamic response of amplified photodetectors, allowing time-dependent photodiode correction factors for spectral radiance histories from shock experiments. We validate correct operation of the modified Caltech pyrometer with actual shock temperature experiments on single-crystal NaCl and MgO and obtain very good agreement with the literature data for these substances. We conclude with a summary of the most essential requirements for error-free calibration of a fiber-optic shock-temperature pyrometer using a high-power coiled tungsten halogen

  16. Dilution and permeation standards for the generation of NO, NO2 and SO2 calibration gas mixtures

    Science.gov (United States)

    Haerri, H.-P.; Macé, T.; Waldén, J.; Pascale, C.; Niederhauser, B.; Wirtz, K.; Stovcik, V.; Sutour, C.; Couette, J.; Waldén, T.

    2017-03-01

    The evaluation results of the metrological performance of a dilution and a permeation standard for generating SI-traceable calibration gas mixtures of NO, SO2 and NO2 for ambient air measurements are presented. The composition of the in situ produced reference gas mixtures is calculated from the instantaneous values of the input quantities of the generating standards. In a measurement comparison, the calibration and measurement capabilities of five laboratories were evaluated for the three analytes at limiting amount of substance fractions in ambient air between 20 and 150 nmol mol-1. For the upper generated reference values the target relative uncertainties of  ⩽2% (for NO and SO2) and  ⩽3% (for NO2) for evaluating the laboratory results were fulfilled in 12 out of 13 cases. For the analytical results seven out of nine laboratories met the criteria for the upper values for NO and NO2, for SO2 it was one out of four. From the negative degrees of equivalence of all NO2 comparison results it was supposed that the permeation rate of NO2 through the FEP polymer membrane of the permeator was different in air and N2. Subsequent precision permeation measurements with various carrier gases revealed that the permeation rate of NO2 was  ≈0.8% lower in synthetic air compared to N2. With the corrected NO2 reference values for air the degrees of equivalence of the laboratory results were improved and closer to be symmetrically distributed.

  17. CRM Assessment: Determining the Generalization of Rater Calibration Training. Summary of Research Report: Gold Standards Training

    Science.gov (United States)

    Baker, David P.

    2002-01-01

    The extent to which pilot instructors are trained to assess crew resource management (CRM) skills accurately during Line-Oriented Flight Training (LOFT) and Line Operational Evaluation (LOE) scenarios is critical. Pilot instructors must make accurate performance ratings to ensure that proper feedback is provided to flight crews and appropriate decisions are made regarding certification to fly the line. Furthermore, the Federal Aviation Administration's (FAA) Advanced Qualification Program (AQP) requires that instructors be trained explicitly to evaluate both technical and CRM performance (i.e., rater training) and also requires that proficiency and standardization of instructors be verified periodically. To address the critical need for effective pilot instructor training, the American Institutes for Research (AIR) reviewed the relevant research on rater training and, based on "best practices" from this research, developed a new strategy for training pilot instructors to assess crew performance. In addition, we explored new statistical techniques for assessing the effectiveness of pilot instructor training. The results of our research are briefly summarized below. This summary is followed by abstracts of articles and book chapters published under this grant.

  18. The ACS survey of globular clusters. XIII. Photometric calibration in comparison with Stetson standards

    CERN Document Server

    Hempel, Maren; Anderson, Jay; Aparicio, Antonio; Bedin, Luigi R; Chaboyer, Brian; Majewski, Steven R; Marín-Franch, Antonio; Milone, Antonino; Paust, Nathaniel E Q; Piotto, Giampaolo; Reid, I Neill; Rosenberg, Alfred; Siegel, Michael

    2013-01-01

    In this study we compare the photometric data of 34 Milky Way globular clusters, observed within the ACS Treasury Program (PI: Ata Sarajedini) with the corresponding ground-based data, provided by the Photometric Standard Field Catalogs of Stetson (2000, 2005). We focus on the transformation between the HST/ACS F606W to V-band and F814W to I-band only. The goal is to assess the validity of the filter transformation equations by Sirianni et al.(2005) with respect to their dependence on metallicity, Horizontal Branch morphology, mass and integrated (V-I) colour of the various globular clusters. Such a dependence is expected due to the fact that the transformation equations are based on the observations of only one globular cluster, i.e., NGC 2419. Surprisingly, the correlation between offset and metallicity is found to be weak, with a low level significance. The correlation between offset and Horizontal Branch structure, as well as total cluster mass is still weaker. Based on the available data we do not find t...

  19. A comparative study between different alternatives to prepare gaseous standards for calibrating UV-Ion Mobility Spectrometers.

    Science.gov (United States)

    Criado-García, Laura; Garrido-Delgado, Rocío; Arce, Lourdes; Valcárcel, Miguel

    2013-07-15

    An UV-Ion Mobility Spectrometer is a simple, rapid, inexpensive instrument widely used in environmental analysis among other fields. The advantageous features of its underlying technology can be of great help towards developing reliable, economical methods for determining gaseous compounds from gaseous, liquid and solid samples. Developing an effective method using UV-Ion Mobility Spectrometry (UV-IMS) to determine volatile analytes entails using appropriate gaseous standards for calibrating the spectrometer. In this work, two home-made sample introduction systems (SISs) and a commercial gas generator were used to obtain such gaseous standards. The first home-made SIS used was a static head-space to measure compounds present in liquid samples and the other home-made system was an exponential dilution set-up to measure compounds present in gaseous samples. Gaseous compounds generated by each method were determined on-line by UV-IMS. Target analytes chosen for this comparative study were ethanol, acetone, benzene, toluene, ethylbenzene and xylene isomers. The different alternatives were acceptable in terms of sensitivity, precision and selectivity.

  20. 石油测井仪器的刻度标准化工作%On the Calibration Standardization for Petroleum Nuclear Logging Tools

    Institute of Scientific and Technical Information of China (English)

    陆大卫; 邱益香; 胡秀妮

    2013-01-01

    This paper reviews the construction of the calibration standards & calibration facilities for petroleum nuclear logging tools in China.Emphasized is the significance of calibration facilities to the research & development of well logging tools and their applications.The standard calibration facilities (primary standards) constructed in Logging Metrology Center of Petroleum Industry in China plays an important role in tracing to the source system for nuclear logging tools.The processes of transference,comparison,calibration,testing,verification and inspection are conducted during the tracing to the source system.The establishment of such conversions is in order to ensure the measuring unit uniformity and the accuracy of log data.This paper is aimed at drawing special attention to calibration standards standardization for logging tools and a unified quantities & units,adapting to the internationalization of China petroleum exploration & development.%回顾了中国石油测井仪器刻度标准装置建设的经过,阐明了标准装置对石油测井仪器在研发、应用等方面的重要意义.对石油工业测井计量站已建成的标准装置在国内石油测井仪器量值统一方面所起到的贡献作了阐述.旨在说明原石油部成立石油工业测井计量站的初衷,强调测井仪器刻度标准对石油行业的重要性,希望能够引起有关部门对测井仪器刻度及标准化工作的高度重视,适应中国石油国际化程度不断提高的形势需要.

  1. A novel strategy for preparing calibration standards for the analysis of plant materials by laser-induced breakdown spectroscopy: A case study with pellets of sugar cane leaves

    Energy Technology Data Exchange (ETDEWEB)

    Silva Gomes, Marcos da [Universidade de São Paulo, Centro de Energia Nuclear na Agricultura (CENA), Laboratório de Química Analítica, Caixa Postal 96, CEP 13416-000, Piracicaba, SP (Brazil); Universidade Federal de São Carlos, Departamento de Química, Rodovia Washington Luiz km 235, São Carlos, SP (Brazil); Gustinelli Arantes de Carvalho, Gabriel [Universidade de São Paulo, Centro de Energia Nuclear na Agricultura (CENA), Laboratório de Química Analítica, Caixa Postal 96, CEP 13416-000, Piracicaba, SP (Brazil); Santos, Dário [Universidade Federal de São Paulo, Departamento de Ciências Exatas e da Terra, Rua Professor Arthur Riedel 275, Diadema, SP (Brazil); Krug, Francisco José, E-mail: fjkrug@cena.usp.br [Universidade de São Paulo, Centro de Energia Nuclear na Agricultura (CENA), Laboratório de Química Analítica, Caixa Postal 96, CEP 13416-000, Piracicaba, SP (Brazil)

    2013-08-01

    Calibration is still a challenging task when dealing with the direct analysis of solids. This is particularly true for laser-induced breakdown spectroscopy (LIBS), and laser ablation inductively coupled plasma optical emission spectrometry/mass spectrometry, when the calibrations are matrix-dependent and/or appropriate certified reference materials are generally not available. Looking at the analysis of plant materials in the form of pressed pellets by LIBS, a new method to overcome and/or minimize this difficulty is proposed by keeping the matrix constant in order to produce matrix-matched calibration pellets. To achieve this goal and to test this novel approach, ground sugar cane leaves were chosen and submitted to acid extractions for obtaining the corresponding blank or a material containing very low concentrations of the analytes. The resulting dried solid material was used either as a blank or a low concentration standard, and also homogeneously mixed with the original plant material at appropriate ratios as well. The corresponding pellets were used as calibration standards and ablated at 30 different sites by applying 25 laser pulses per site with a Q-switched Nd:YAG at 1064 nm. The plasma emission collected by lenses was directed through an optical fiber towards a spectrometer equipped with Echelle optics and intensified charge-coupled device. Delay time and integration time gate were fixed at 2.0 and 5.0 μs, respectively. This calibration strategy was tested for the determination of Ca, Mg, K, P, Cu, Mn, and Zn by LIBS in pellets of leaves from 17 varieties of sugar cane and good correlations were obtained with inductively coupled plasma optical emission spectrometry results in the corresponding acid digests. The proposed approach was also useful to estimate the limits of detection based on measurements of blanks, as recommended by IUPAC, or with the aid of a low concentration standard. - Highlights: • Blanks and/or low concentration standards of plant

  2. SECONDARY STANDARD CALIBRATION, MEASUREMENT AND IRRADIATION CAPABILITIES OF THE INDIVIDUAL MONITORING SERVICE AT THE HELMHOLTZ ZENTRUM MÜNCHEN: ASPECTS OF UNCERTAINTY AND AUTOMATION.

    Science.gov (United States)

    Greiter, M B; Denk, J; Hoedlmoser, H

    2016-09-01

    The individual monitoring service at the Helmholtz Zentrum München has adopted the recommendations of the ISO 4037 and 6980 standards series as base of its dosimetric systems for X-ray, gamma and beta dosimetry. These standards define technical requirements for radiation spectra and measurement processes, but leave flexibility in the implementation of irradiations as well as in the resulting uncertainty in dose or dose rate. This article provides an example for their practical implementation in the Munich IAEA/WHO secondary standard dosimetry laboratory. It focusses on two aspects: automation issues and uncertainties in calibration.

  3. Using standard calibrated geometries to characterize a coaxial high purity germanium gamma detector for Monte Carlo simulations

    NARCIS (Netherlands)

    van der Graaf, E. R.; Dendooven, P.; Brandenburg, S.

    2014-01-01

    A detector model optimization procedure based on matching Monte Carlo simulations with measurements for two experimentally calibrated sample geometries which are frequently used in radioactivity measurement laboratories results in relative agreement within 5% between simulated and measured efficienc

  4. 论衡器检定工作中的规范化%On Standardization of Calibration of Weighing Machines

    Institute of Scientific and Technical Information of China (English)

    康峰

    2011-01-01

    This paper dissertates the definition of compulsory testing on calibration of weighing machines according to the national regulation, problems and considerations during the prosess of calibration of weighing machines.%本文论述了根据国家检定规程开展衡器检定工作中衡器强检界定,衡器检定中存在的问题及衡器检定中的注意事项。

  5. 基于标准球的显微图像测量系统标定方法%Micro-Image Measuring System Calibration Based on Standard Ball

    Institute of Scientific and Technical Information of China (English)

    聂凯; 刘文耀; 王晋疆

    2013-01-01

    A camera calibration method using standard ball is proposed for micro-image measuring system that has small field of view and small depth of field. The method simplified the computational modeling and reduced calibration parameters. The scale factor can be calculated by means of the ratio of horizontal diameter to vertical diameter of the image of standard ball. According to the point set of silhouettes of standard ball, the distortion coefficient and the position of principal point of imaging system can be calculated by optimization method. The magnification of micro imaging system can be figured out based on the actual diameter of standard ball. Using the feature that the contours of standard ball placed in the common field of view of multiple cameras are always visible from any position,multiple cameras in micro imaging system can be calibrated simultaneously, so the process of calibration will be simplified. The experimental results show that calibration precision is high and the limit error, 3a, of the system calibrated by this method is 2.41 μm.%提出一种利用标准球对显微图像测量系统进行标定的方法.该方法利用显微系统小视场小景深的成像特点简化计算模型,减少标定参数.通过标准球图像在水平和垂直方向上的直径比计算比例因子;利用标准球边缘图像的边缘点集,运用优化的方法来计算成像系统的畸变系数和主点位置.系统的放大倍数由标准球的实际直径来标定得出.利用标准球在多摄像机公共视场内其轮廓在任何位置均可见这一特性,可同时对显微图像测量系统中的多个摄像机进行标定,简化标定过程.实验结果表明,该方法标定精度较高,标定后的测量系统的极限误差3σ为2.41μm.

  6. Air kerma standardization for diagnostic radiology, and requirements proposal for calibration laboratories; Padronizacao da grandeza Kerma no ar para radiodiagnostico e proposta de requisitos para laboratorios de calibracao

    Energy Technology Data Exchange (ETDEWEB)

    Ramos, Manoel Mattos Oliveira

    2009-07-01

    The demand for calibration services and quality control in diagnostic radiology has grown in the country since the publication of the governmental regulation 453, issued by the Ministry of Health in 1998. At that time, to produce results facing the new legislation, many laboratories used different standards and radiation qualities, some of which could be inadequate. The international standards neither supplied consistent radiation qualities and standardization for the different types of equipment available. This situation changed with the publication of the new edition of the IEC 61267 standard, published in 2005. A metrology network was created, but it is not yet accredited by the accreditation organism of the country, INMETRO. The objective of this work was to implement the standardization of the air kerma for the un attenuated qualities (RQR) of IEC 61267, and to develop a requirement proposal for instruments calibration laboratories. Results of interlaboratory comparisons demonstrate that the quantity is standardized and internationally traceable. A laboratory requirement proposal was finalized and it shall be submitted to INMETRO to be used as auxiliary normative document in laboratory accreditation. (author)

  7. Standardization of 64Cu and 68Ga by the 4π(PC)β-γ coincidence method and calibration of the ionization chamber.

    Science.gov (United States)

    Sahagia, M; Luca, A; Antohe, A; Ivan, C

    2012-09-01

    The paper treats the application of the 4π(PC)β-γ coincidence method for the standardization of the radionuclides (64)Cu and (68)Ga. The general coincidence equations are written. Two types of extrapolation were described and used in measurement: the positron-annihilation coincidence, and the counting of all emitted radiations; both methods are compared with respect to results, advantages and drawbacks. The impurities' content correction was applied. The standardized solutions were used to calibrate the ionization chamber CENTRONIC IG12/20A and to determine the gamma-rays emission intensities.

  8. Calibration of the modulation transfer function of surface profilometers with binary pseudo-random test standards: expanding the application range

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, Valeriy V.; Anderson, Erik H.; Barber, Samuel K.; Bouet, Nathalie; Cambie, Rossana; Conley, Raymond; McKinney, Wayne R.; Takacs, Peter Z.; Voronov, Dmitriy L.

    2011-03-14

    A modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) gratings and arrays [Proc. SPIE 7077-7 (2007), Opt. Eng. 47, 073602 (2008)] has been proven to be an effective MTF calibration method for a number of interferometric microscopes and a scatterometer [Nucl. Instr. and Meth. A616, 172 (2010)]. Here we report on a further expansion of the application range of the method. We describe the MTF calibration of a 6 inch phase shifting Fizeau interferometer. Beyond providing a direct measurement of the interferometer's MTF, tests with a BPR array surface have revealed an asymmetry in the instrument's data processing algorithm that fundamentally limits its bandwidth. Moreover, the tests have illustrated the effects of the instrument's detrending and filtering procedures on power spectral density measurements. The details of the development of a BPR test sample suitable for calibration of scanning and transmission electron microscopes are also presented. Such a test sample is realized as a multilayer structure with the layer thicknesses of two materials corresponding to BPR sequence. The investigations confirm the universal character of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  9. Update On The Low Background IR Calibration Facility At The National Institute Of Standards And Technology (Formerly NBS)

    Science.gov (United States)

    Ebner, Stephen C.; Parr, Albert C.; Hoyt, Clifford C.

    1989-09-01

    Details will be given about the recently completed facility for the calibration of infrared sources in a low background environment. The basic components of the facility are a large (60cm diameter by 152cm long) stainless steel vacuum chamber housed in a soft-wall cleanroom. A low background environment inside the chamber is achieved by cooling internal cryoshields to temperatures less than 20K using a closed cycle helium refrigerator. Sources of up to 30cm on a side can be inserted into the chamber for calibration. Total radiant power from the blackbodies is measured with an Absolute Cryogenic Radiometer. Plans will be discussed for future enhancement of the system allowing for measurement of spectral and angular distribution of the emitted radiation and possible experiments which could utilize the full capabilities of this system.

  10. In-situ calibration of clinical built-in KAP meters with traceability to a primary standard using a reference KAP meter

    Science.gov (United States)

    Malusek, A.; Helmrot, E.; Sandborg, M.; Grindborg, J.-E.; Alm Carlsson, G.

    2014-12-01

    The air kerma-area product (KAP) is used for settings of diagnostic reference levels. The International Atomic Energy Agency (IAEA) recommends that doses in diagnostic radiology (including the KAP values) be estimated with an accuracy of at least ±7% (k = 2). Industry standards defined by the International Electrotechnical Commission (IEC) specify that the uncertainty of KAP meter measurements should be less than +/- 25% (k = 2). Medical physicists willing to comply with the IAEA’s recommendation need to apply correction factors to KAP values reported by x-ray units. The aim of this work is to present and evaluate a calibration method for built-in KAP meters on clinical x-ray units. The method is based on (i) a tandem calibration method, which uses a reference KAP meter calibrated to measure the incident radiation, (ii) measurements using an energy-independent ionization chamber to correct for the energy dependence of the reference KAP meter, and (iii) Monte Carlo simulations of the beam quality correction factors that correct for differences between beam qualities at a standard laboratory and the clinic. The method was applied to the KAP meter in a Siemens Aristos FX plus unit. It was found that values reported by the built-in KAP meter differed from the more accurate values measured by the reference KAP meter by more than 25% for high tube voltages (more than 140 kV) and heavily filtered beams (0.3 mm Cu). Associated uncertainties were too high to claim that the IEC’s limit of 25% was exceeded. Nevertheless the differences were high enough to justify the need for a more accurate calibration of built-in KAP meters.

  11. SEALDH-II: An airborne, autonomous, calibration-free TDLAS Hygrometer. First in-flight results with metrological links to the German Primary Humidity Standards

    Science.gov (United States)

    Buchholz, B.; Ebert, V.

    2014-12-01

    The accuracy and comparability of airborne hygrometers remains under debate often only reaching ±10% under static lab conditions or even >±30% in flight. This often limits atmospheric models as well as other species sensors which need to be corrected for water vapor dilution or cross sensitivity. Further challenges come from the huge dynamic range (2-40000 ppmv), the strong spatial gradients (up to several 1000 ppmv/s). and the difficult and error-prone calibration processes, very often without any direct links to the accurate global metrological water vapor scales, which are defined via national primary humidity generators. To fill this gap and provide "traceable", i.e. metrologically validated, airborne hygrometers we developed the Selective Extractive Airborne Laser Diode Hygrometer, SEALDH, which uses dTDLAS with a special, calibration-free data evaluation to circumvent the need for frequent field calibrations and to ensure high accuracy and comparability by a holistic data quality assurance concept in combination with extensive metrological validations at national primary standards. Thus SEALDH-II can be used in a calibration-free field sensor mode (with an absolute, metrologically defined uncertainty of 4.3% +- 3ppmv). The response time is mainly limited by the gas flow and significantly below 1 sec, yielding precision down to 0.08 ppmv (1σ, 1sec) measured at 600 ppmv and 1000 hPa. Its excellent long-term stability, field transfer standard. SEALDH-II has been recently operated without any failures for over 50 hours on several airborne science missions (DENCHAR, AIRTOSS-I, and AIRTOSS-II). In addition SEALDH participated in the representative, international comparison AquaVIT-II. The performance during these missions and the value of a traceability will be demonstrated and discussed.

  12. Measuring laser carrier-envelope phase effects in the noble gases with an atomic hydrogen calibration standard

    CERN Document Server

    Khurmi, Champak; U, Satya Sainadh; Ivanov, I A; Kheifets, A S; Tong, X M; Litvinyuk, I V; Sang, R T; Kielpinski, D

    2016-01-01

    We present accurate measurements of carrier-envelope phase effects on ionisation of the noble gases with few-cycle laser pulses. The experimental apparatus is calibrated by using atomic hydrogen data to remove any systematic offsets and thereby obtain accurate CEP data on other generally used noble gases such as Ar, Kr and Xe. Experimental results for H are well supported by exact TDSE theoretical simulations however significant differences are observed in case of noble gases.

  13. (60)Co in cast steel matrix: A European interlaboratory comparison for the characterisation of new activity standards for calibration of gamma-ray spectrometers in metallurgy.

    Science.gov (United States)

    Tzika, Faidra; Burda, Oleksiy; Hult, Mikael; Arnold, Dirk; Marroyo, Belén Caro; Dryák, Pavel; Fazio, Aldo; Ferreux, Laurent; García-Toraño, Eduardo; Javornik, Andrej; Klemola, Seppo; Luca, Aurelian; Moser, Hannah; Nečemer, Marijan; Peyrés, Virginia; Reis, Mario; Silva, Lidia; Šolc, Jaroslav; Svec, Anton; Tyminski, Zbigniew; Vodenik, Branko; Wätjen, Uwe

    2016-08-01

    Two series of activity standards of (60)Co in cast steel matrix, developed for the calibration of gamma-ray spectrometry systems in the metallurgical sector, were characterised using a European interlaboratory comparison among twelve National Metrology Institutes and one international organisation. The first standard, consisting of 14 disc shaped samples, was cast from steel contaminated during production ("originally"), and the second, consisting of 15 similar discs, from artificially-contaminated ("spiked") steel. The reference activity concentrations of (60)Co in the cast steel standards were (1.077±0.019) Bqg(-1) on 1 January 2013 12h00 UT and (1.483±0.022) Bqg(-1) on 1 June 2013 12h00 UT, respectively.

  14. X射线环境水平标准装置的研究与建立%The establishment of x-ray environmental standard calibration quality

    Institute of Scientific and Technical Information of China (English)

    欧向明; 赵士庵; 丁艳秋

    2014-01-01

    Objective:According to the requirement of ISO and National Standard, new X-ray environmental standard for the calibration of environmental dosimeter was established in Beijing SSDL. Methods:Half value layer and air kerma dose-rate was determined in the X-ray environmental qualities by a standard dosimeter. Results: The technical requirement of ISO and National Standard are met within 2% for the X-ray environmental standard qualities in Beijing SSDL. Conclusion:The X-ray standard radiation field can be used for the calibration of environmental dosimeters and experiment of research.%目的:按照国际标准化组织(ISO)技术规范及我国现行国家标准的相关要求,在国家二级标准剂量学实验室建立用于校准X射线环境水平剂量仪的标准装置。方法:通过标准剂量仪测量不同附加过滤后的X射线束的衰减,测定标准辐射场常用的半值层及空气比释动能率,确定环境水平X射线辐射线质的有效能量和平均能量。结果:测定的辐射线质等参数与ISO和国家标准给出的技术指标误差<2%,满足标准装置的要求。结论:建立的X射线环境水平标准装置可以用于校准环境水平辐射剂量仪、并为开展相关的研究工作提供实验条件。

  15. Fabrication of grating-like polystyrene latex monolayer structure as three-dimensional calibration standards for scanning probe microscope

    Institute of Scientific and Technical Information of China (English)

    Zhu Guo-Dong; Zeng Zhi-Gang; Guo Zhang; Du Qiang-Guo; Yan Xue-Jian

    2009-01-01

    This paper illuminates the preparation of grating-like polystyrene latex monolayer structure, which can minimize the effects of the size deviation of spheres and the defect transfer on the accuracy as calibration samples for micro-scopes. The latex films are grown on freshly cleaved mica substrates by vertical deposition method. The concentration dependence of the structure and the topography of latex films is characterized by optical microscope, ultraviolet-visible transmission spectrum and scanning probe microscope. The origination of such a grating-like structure is also discussed.

  16. Electroluminescent TCC, C3dg and fB/Bb epitope assays for profiling complement cascade activation in vitro using an activated complement serum calibration standard.

    Science.gov (United States)

    van Vuuren, B Jansen; Bergseth, G; Mollnes, T E; Shaw, A M

    2014-01-15

    Electroluminescent assays for epitopes on the complement components C3dg, terminal complement complex (TCC) and factor B/Bb (fB/Bb) have been developed with capture and detection antibodies to produce detection limits C3dg=91±9ng/mL, TCC=3±0.1ng/mL and fB=55.7±0.1ng/mL. The assay performance was assessed against a series of zymosan and heat aggregated IgG (HAIgG) in vitro activations of complement using a calibrated activated complement serum (ACS) as calibration standard. The ACS standard was stable within 20% accuracy over a 6-month period with freeze-thaw cycles as required. Differential activation of the complement cascade was observed for TCC showing a pseudo-first order formation half-life of 3.5h after activation with zymosan. The C3dg activation fragment indicates a 10% total activation for both activation agents. The kinetic-epitope analysis for fB indicates that the capture epitope is on the fB/Bb protein fragment which can then become covered by the formation of C3bBb or C3bBbP complexes during the time course of the cascade.

  17. Final report on key comparison CCAUV.A-K5: pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 kHz

    Science.gov (United States)

    Avison, Janine; Barham, Richard

    2014-01-01

    This document and the accompanying spreadsheets constitute the final report for key comparison CCAUV.A-K5 on the pressure calibration of laboratory standard microphones in the frequency range from 2 Hz to 10 kHz. Twelve national measurement institutes took part in the key comparison and the National Physical Laboratory piloted the project. Two laboratory standard microphones IEC type LS1P were circulated to the participants and results in the form of regular calibration certificates were collected throughout the project. One of the microphones was subsequently deemed to have compromised stability for the purpose of deriving a reference value. Consequently the key comparison reference value (KCRV) has been made based on the weighted mean results for sensitivity level and for sensitivity phase from just one of the microphones. Corresponding degrees of equivalence (DoEs) have also been calculated and are presented. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  18. Lidar to lidar calibration

    DEFF Research Database (Denmark)

    Fernandez Garcia, Sergio; Villanueva, Héctor

    This report presents the result of the lidar to lidar calibration performed for ground-based lidar. Calibration is here understood as the establishment of a relation between the reference lidar wind speed measurements with measurement uncertainties provided by measurement standard and correspondi...

  19. Preparation of Calibration Standards of N1-H Paralytic Shellfish Toxin Analogues by Large-Scale Culture of Cyanobacterium Anabaena circinalis (TA04

    Directory of Open Access Journals (Sweden)

    Toshiyuki Suzuki

    2011-03-01

    Full Text Available Mouse bioassay is the official testing method to quantify paralytic shellfish toxins (PSTs in bivalves. A number of alternative analytical methods have been reported. Some methods have been evaluated by a single laboratory validation. Among the different types of methods, chemical analyses are capable of identifying and quantifying the toxins, however a shortage of the necessary calibration standards hampers implementation of the chemical analyses in routine monitoring of PSTs in bivalves. In our present study, we studied preparation of major PST analogues as calibrants by large-scale cultivation of toxic freshwater cyanobacteria Anabaena circinalis TA04. The cells were steadily grown in 10 L bottle for 28 days. The primary N1-H toxins, C1/C2, were produced at a concentration of 1.3 ± 0.1 µmol/L. The intracellular and extracellular toxins occupied 80% and 20%, respectively. Over 220 µmol of the toxins was obtained from approximately 200 L of the culture over six months, demonstrating that it is sufficient to prepare saxitoxin analogues. The toxins were chemically converted to six N1-H analogues. Preparation of the analogues was carried out at relatively high yields (50–90%. The results indicate that our preparation method is useful to produce N1-H toxins. In our present study, detailed conditions for preparation of one of the rare N1-H analogues, gonyautoxin-5, were investigated.

  20. The Science of Calibration

    Science.gov (United States)

    Kent, S. M.

    2016-05-01

    This paper presents a broad overview of the many issues involved in calibrating astronomical data, covering the full electromagnetic spectrum from radio waves to gamma rays, and considering both ground-based and space-based missions. These issues include the science drivers for absolute and relative calibration, the physics behind calibration and the mechanisms used to transfer it from the laboratory to an astronomical source, the need for networks of calibrated astronomical standards, and some of the challenges faced by large surveys and missions.

  1. The PROMIS Physical Function item bank was calibrated to a standardized metric and shown to improve measurement efficiency

    DEFF Research Database (Denmark)

    Rose, Matthias; Bjørner, Jakob; Gandek, Barbara;

    2014-01-01

    OBJECTIVE: To document the development and psychometric evaluation of the Patient-Reported Outcomes Measurement Information System (PROMIS) Physical Function (PF) item bank and static instruments. STUDY DESIGN AND SETTING: The items were evaluated using qualitative and quantitative methods. A total...... of 16,065 adults answered item subsets (n>2,200/item) on the Internet, with oversampling of the chronically ill. Classical test and item response theory methods were used to evaluate 149 PROMIS PF items plus 10 Short Form-36 and 20 Health Assessment Questionnaire-Disability Index items. A graded...... response model was used to estimate item parameters, which were normed to a mean of 50 (standard deviation [SD]=10) in a US general population sample. RESULTS: The final bank consists of 124 PROMIS items covering upper, central, and lower extremity functions and instrumental activities of daily living...

  2. Development and validation of a SYBR-Green I real-time PCR protocol to detect hazelnut (Corylus avellana L.) in foods through calibration via plasmid reference standard.

    Science.gov (United States)

    D'Andrea, Matteo; Coïsson, Jean Daniel; Travaglia, Fabiano; Garino, Cristiano; Arlorio, Marco

    2009-12-09

    Many tree nuts are considered to be a serious problem in food safety, because of the presence of causative factors in IgE-mediated food allergies. Among these, hazelnut (Corylus avellana L.) seeds are largely used in a range of confectionery products and contain many well-characterized allergens. DNA-based methods and ELISA tests may prove to be useful to assess the presence of hidden ingredients in foods. The aim of this work was the development and validation of a species-specific SYBR Green I real-time PCR protocol for the detection of hazelnut in foods. A novel efficient primer pair on the Cor a 8 genomic coding region was designed by preparing a plasmid vector-based internal reference standard to calibrate the PCR. A good sensitivity, down to 20 (genomic) and 15 (plasmid) DNA copies, was established. All of the commercial samples considered in our study (containing hazelnut as ingredient or as a potential trace cross-contamination) were effectively amplified by PCR, showing a perfect correspondence with an ELISA commercial test, employed as a reference standard method.

  3. Problems in Preparation and Calibration of Concentration of EDTA Standard Solution%配制和标定EDTA标准溶液浓度时要注意的问题

    Institute of Scientific and Technical Information of China (English)

    叶萍萍

    2011-01-01

    本文主要从配制和标定EDTA标准溶液浓度入手,分析其过程中应该注意的几个问题,从而标定出EDTA标准溶液的准确浓度.%This article begins with the preparation and calibration of the concentration of EDTA standard solution, analyzes the problems in the process which should pay attention to in order to calibrate the accurate concentration of EDTA standard solution.

  4. Sandia WIPP calibration traceability

    Energy Technology Data Exchange (ETDEWEB)

    Schuhen, M.D. [Sandia National Labs., Albuquerque, NM (United States); Dean, T.A. [RE/SPEC, Inc., Albuquerque, NM (United States)

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities.

  5. Flexible calibration procedure for fringe projection profilometry

    OpenAIRE

    Vargas, Javier; Quiroga Mellado, Juan Antonio; Terrón López, María José

    2007-01-01

    A novel calibration method for whole field three-dimensional shape measurement by means of fringe projection is presented. Standard calibration techniques, polynomial-and model-based, have practical limitations such as the difficulty of measuring large fields of view, the need to use precise z stages, and bad calibration results due to inaccurate calibration points. The proposed calibration procedure is a mixture of the two main standard techniques, sharing their benefits and avoiding their m...

  6. Determination of As, Cd, Pb and Tl in coal by electrothermal vaporization inductively coupled plasma mass spectrometry using slurry sampling and external calibration against aqueous standards

    Energy Technology Data Exchange (ETDEWEB)

    Borges, D.L.G.; Welz, B.; Curtius, A.J. [University of Federal Santa Catarina, Florianopolis (Brazil)

    2007-06-15

    A method was developed for the determination of As, Cd, Pb and Tl by electrothermal vaporization inductively coupled plasma mass spectrometry (ETV-ICP-MS) with slurry sample introduction using certified reference coal samples. The As-75, Cd-111, Pb-208 and Tl-205 isotopes were monitored, considering the lower probabilities of these isotopes in suffering interferences. The carrier and modifier effect of Ru added in solution was evaluated, demonstrating that sensitivity for all elements is considerably improved by the addition of 15 mg of Ru to each individual measurement. This confirms its ability to act as a physical carrier particularly for the analytes in aqueous solution. After optimization of the operational parameters, the determination of the four elements in six certified reference coal samples was carried out by external calibration against aqueous standards in 5% v= v HNO{sub 3}, resulting in good agreement between the certified or given values and the determined ones. For Tl, due to the absence of certified values, a comparison was established considering previously published data using solid sampling high-resolution continuum source graphite furnace atomic absorption spectrometry. Detection limits ({mu} g g{sup -1}) of 0.1, 0.004, 0.045 and 0.001 were achieved for As, Cd, Pb and Tl, respectively, and the precision was typically better than 10%.

  7. Automatic Titrator Calibration Standard Solution of Silver Nitrate%自动电位滴定仪标定硝酸银标准溶液

    Institute of Scientific and Technical Information of China (English)

    马玉芬

    2012-01-01

      Using the configuration DM141-SC-type silver electrode DL58 automatic titrator calibration standard solution of silver nitrate, accuracy and precision compared with manual titration with a large degree of increase, and simple operation, equipment maintenance, more accurate titration results .Titration process signal acquisition is 0.1mV, minimum feed rate can be titrated to achieve 0.001ml, is the recommended method.%  采用配置DM141-SC型银电极的DL58型自动电位滴定仪标定硝酸银标准溶液,准确度及精密度均较手动滴定有很大程度的提高,且操作简单,仪器维护方便,滴定结果更准确,滴定过程采集信号为0.1mV,滴定最小进给量可达到0.001mL,是目前值得推荐的方法.

  8. Calibration of surface roughness standards

    DEFF Research Database (Denmark)

    Thalmann, R.; Nicolet, A.; Meli, F.

    2016-01-01

    The key comparison EURAMET.L-K8.2013 on roughness was carried out in the framework of a EURAMET project starting in 2013 and ending in 2015. It involved the participation of 17 National Metrology Institutes from Europe, Asia, South America and Africa representing four regional metrology organisat...

  9. Lidar to lidar calibration

    DEFF Research Database (Denmark)

    Georgieva Yankova, Ginka; Courtney, Michael

    This report presents the result of the lidar to lidar calibration performed for ground-based lidar. Calibration is here understood as the establishment of a relation between the reference lidar wind speed measurements with measurement uncertainties provided by measurement standard and corresponding...... lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from the reference lidar measurements are given for information only....

  10. Site Calibration report

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Vesth, Allan

    This report describes the site calibration carried out at Østerild, during a given period. The site calibration was performed with two Windcube WLS7 (v1) lidars at ten measurements heights. The lidar is not a sensor approved by the current version of the IEC 61400-12-1 [1] and therefore the site...... calibration with lidars does not comply with the standard. However, the measurements are carried out following the guidelines of IEC 61400-12-1 where possible, but with some deviations presented in the following chapters....

  11. Standardization of calibration of clinic dosemeters using electric currents and charges; Padronizacao da calibracao de dosimetros clinicos utilizando cargas e correntes eletricas

    Energy Technology Data Exchange (ETDEWEB)

    Peres, Marcos Antonio de Lima

    1999-09-15

    Clinical dosimeters used in radiotherapy are calibrated in Latin American countries, including Brazil, as a complete 'system, i.e., ionization chamber and electrometer together. Some countries, as European ones, and USA calibrate them by component, i.e., ionisation chamber apart from electrometer. This method is more advantageous than the previous one, since it makes possible the automation of the calibration process, allowing the acquisition of data related to the chamber and the electrometer measurements independently, as well as the substitution of one of the components, in case of failure. This work proposes a method for standardisation of low intensity electric charge and currents in order to implement electric calibration of electrometers. This will make possible the implementation of a by components calibration technique, by components, of clinical dosimeters in Brazil. The results obtained with the calibration by components prove that the proposed method of standardisation of low electric charges and currents is viable. The difficulties found for the generation and measurement of low intensity electric charges and currents and the procedures adopted for the calibration by components are presented. Additionally, a current source was built to yield reference electric charges that will make possible the quality control clinical dosimeters. This source will also allow the user to identify the defective components of the dosimeters, through a simple verification test. (author)

  12. Primary Research on the Standard System of Air Vehicle Calibration Models Used in Wind Tunnel Test%飞行器风洞试验标模体系研究初探

    Institute of Scientific and Technical Information of China (English)

    战培国; 罗月培

    2011-01-01

    风洞试验标模是一种评估风洞试验准度和验证CFD算法的通用校准模型.本文归纳分析了北大西洋公约组织AGARD系列、法国ONERAM系列为代表的国外风洞试验标模发展情况,阐述了我国DBM、GBM、HSCM风洞标模系列的发展现状,探讨了建立和完善风洞试验标模体系的一些问题,意在为国内风洞和试验技术发展提供参考.%The standard air vehicle calibration model is a common calibration model used for evaluating wind tunnel test results and validating CFD methods.This paper analyses the development of overseas wind tunnel calibration models,typically,NATO AGARD series and ONERA M series in France,introduces state of the art domestic calibration models,such as DBM,CBM and HSCM series,discusses some problems about setting up and consummating the system of calibration models.The purpose is to provide thoughts and references for the development of domestic wind tunnel techniques.

  13. 基于ASAM标准的汽车电控系统匹配标定系统设计%Design of Automotive Electronic Control System Calibration System based on ASAM Standards

    Institute of Scientific and Technical Information of China (English)

    温泉; 张广秀; 张建

    2012-01-01

    ASAM-MCD standards is a set of widely-agreed calibration criterion in industry. Application of the criterion can standardize data exchange methods, software and hardware tools development, and thereby decrease development cost and periods. This paper discusses the main features of CCP and ASAP2 standardized interface for the description data. We have developed the electric vehicle control unit calibration system according to the ASAM-MCD standards. The system can implement file-based parameters monitoring, measurement and calibration.%ASAM—MCD标准是一套得到工业界广泛认可的匹配标定规范。通过该规范的实施,可以对ECU开发过程中用到的数据交换方法和软硬件工具进行标准化,从而减少汽车电子领域的开发成本、缩短开发周期。本文论述了ASAM—MCD标准中的CAN标定协议和ASAP2数据接口规范的主要特点,并根据该标准开发了电动汽车用整车控制器匹配标定系统。通过该系统可以实现基于文件的参数监控、测量和标定。

  14. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  15. Internet-based calibration of a multifunction calibrator

    Energy Technology Data Exchange (ETDEWEB)

    BUNTING BACA,LISA A.; DUDA JR.,LEONARD E.; WALKER,RUSSELL M.; OLDHAM,NILE; PARKER,MARK

    2000-04-17

    A new way of providing calibration services is evolving which employs the Internet to expand present capabilities and make the calibration process more interactive. Sandia National Laboratories and the National Institute of Standards and Technology are collaborating to set up and demonstrate a remote calibration of multifunction calibrators using this Internet-based technique that is becoming known as e-calibration. This paper describes the measurement philosophy and the Internet resources that can provide real-time audio/video/data exchange, consultation and training, as well as web-accessible test procedures, software and calibration reports. The communication system utilizes commercial hardware and software that should be easy to integrate into most calibration laboratories.

  16. Calibration Database Logic Analysis and Implementation based on ASAM Standard%基于ASAM标准的标定数据库逻辑分析与编程实现

    Institute of Scientific and Technical Information of China (English)

    宋维群; 杨世春; 李明; 崔海港; 曹耀光

    2012-01-01

    In this article, ASAM -2MC standard proposed by ASAM organization standardizes data needed in calibration of automotive electronic control system. In-depth analysis is made to ASAM-2MC standard and logic relation and database algorithm between calibrated information items are introduced, and calibration database management software in compliance with ASAM-2MC standard is developed. The interpretation of a21 description document used in calibration of engine electronic control system shows that this software is able to classify control parameters in vehicle development, display and provide visual interface to analyze and correct the items in the interface.%ASAM组织提出的ASAM-2MC标准对汽车电子控制系统标定过程中所需要的数据信息进行了规范.深入分析了ASAM-2MC标准并提出了相应的标定信息项目实体之间的逻辑关系及数据库解析算法,开发了符合ASAM-2MC标准的标定数据库管理软件.通过对发动机电控系统标定过程中使用的a21描述文件的解析表明,该软件能够对整车开发过程中与控制相关的参数信息进行分类,显示并提供可视化界面对其中的项目进行分析与修改.

  17. NVLAP calibration laboratory program

    Energy Technology Data Exchange (ETDEWEB)

    Cigler, J.L.

    1993-12-31

    This paper presents an overview of the progress up to April 1993 in the development of the Calibration Laboratories Accreditation Program within the framework of the National Voluntary Laboratory Accreditation Program (NVLAP) at the National Institute of Standards and Technology (NIST).

  18. 14 CFR 33.45 - Calibration tests.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 1 2010-01-01 2010-01-01 false Calibration tests. 33.45 Section 33.45... STANDARDS: AIRCRAFT ENGINES Block Tests; Reciprocating Aircraft Engines § 33.45 Calibration tests. (a) Each engine must be subjected to the calibration tests necessary to establish its power characteristics...

  19. 14 CFR 33.85 - Calibration tests.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 1 2010-01-01 2010-01-01 false Calibration tests. 33.85 Section 33.85... STANDARDS: AIRCRAFT ENGINES Block Tests; Turbine Aircraft Engines § 33.85 Calibration tests. (a) Each engine must be subjected to those calibration tests necessary to establish its power characteristics and...

  20. A new experimental procedure for determination of photoelectric efficiency of a NaI(Tl) detector used for nuclear medicine liquid waste monitoring with traceability to a reference standard radionuclide calibrator.

    Science.gov (United States)

    Ceccatelli, A; Campanella, F; Ciofetta, G; Marracino, F M; Cannatà, V

    2010-02-01

    To determine photopeak efficiency for (99m)Tc of the NaI(Tl) detector used for liquid waste monitoring at the Nuclear Medicine Unit of IRCCS Paediatric Hospital Bambino Gesù in Rome, a specific experimental procedure, with traceability to primary standards, was developed. Working with the Italian National Institute for Occupational Prevention and Safety, two different calibration source geometries were employed and the detector response dependence on geometry was investigated. The large percentage difference (almost 40%) between the two efficiency values obtained showed that geometrical effects cannot be neglected.

  1. Mercury Calibration System

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Joseph Rovani; Mark Sanderson; Ryan Boysen; William Schuster

    2009-03-11

    U.S. Environmental Protection Agency (EPA) Performance Specification 12 in the Clean Air Mercury Rule (CAMR) states that a mercury CEM must be calibrated with National Institute for Standards and Technology (NIST)-traceable standards. In early 2009, a NIST traceable standard for elemental mercury CEM calibration still does not exist. Despite the vacature of CAMR by a Federal appeals court in early 2008, a NIST traceable standard is still needed for whatever regulation is implemented in the future. Thermo Fisher is a major vendor providing complete integrated mercury continuous emissions monitoring (CEM) systems to the industry. WRI is participating with EPA, EPRI, NIST, and Thermo Fisher towards the development of the criteria that will be used in the traceability protocols to be issued by EPA. An initial draft of an elemental mercury calibration traceability protocol was distributed for comment to the participating research groups and vendors on a limited basis in early May 2007. In August 2007, EPA issued an interim traceability protocol for elemental mercury calibrators. Various working drafts of the new interim traceability protocols were distributed in late 2008 and early 2009 to participants in the Mercury Standards Working Committee project. The protocols include sections on qualification and certification. The qualification section describes in general terms tests that must be conducted by the calibrator vendors to demonstrate that their calibration equipment meets the minimum requirements to be established by EPA for use in CAMR monitoring. Variables to be examined include linearity, ambient temperature, back pressure, ambient pressure, line voltage, and effects of shipping. None of the procedures were described in detail in the draft interim documents; however they describe what EPA would like to eventually develop. WRI is providing the data and results to EPA for use in developing revised experimental procedures and realistic acceptance criteria based on

  2. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John F. Schabron; Joseph F. Rovani; Susan S. Sorini

    2007-03-31

    The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005, requires that calibration of mercury continuous emissions monitors (CEMs) be performed with NIST-traceable standards. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The traceability protocol will be written by EPA. Traceability will be based on the actual analysis of the output of each calibration unit at several concentration levels ranging from about 2-40 ug/m{sup 3}, and this analysis will be directly traceable to analyses by NIST using isotope dilution inductively coupled plasma/mass spectrometry (ID ICP/MS) through a chain of analyses linking the calibration unit in the power plant to the NIST ID ICP/MS. Prior to this project, NIST did not provide a recommended mercury vapor pressure equation or list mercury vapor pressure in its vapor pressure database. The NIST Physical and Chemical Properties Division in Boulder, Colorado was subcontracted under this project to study the issue in detail and to recommend a mercury vapor pressure equation that the vendors of mercury vapor pressure calibration units can use to calculate the elemental mercury vapor concentration in an equilibrium chamber at a particular temperature. As part of this study, a preliminary evaluation of calibration units from five vendors was made. The work was performed by NIST in Gaithersburg, MD and Joe Rovani from WRI who traveled to NIST as a Visiting Scientist.

  3. An empirical calibration for {sup 4}He quantification in minerals and rocks by laser fusion and noble gas mass spectrometry using Cerro de Mercado (Durango, Mexico) fluorapatite as a standard

    Energy Technology Data Exchange (ETDEWEB)

    Sole, Jesus [Instituto de Geologia, Universidad Nacional Autonoma de Mexico, Cd. Universitaria, 04510 Mexico, D.F. (Mexico)]. E-mail: jsole@geologia.unam.mx; Pi, Teresa [Instituto de Geologia, Universidad Nacional Autonoma de Mexico, Cd. Universitaria, 04510 Mexico, D.F. (Mexico)

    2005-04-11

    An empirical calibration with a natural mineral standard (fluorapatite from Cerro de Mercado, Mexico) is proposed as a method to determine the {sup 4}He concentration of mineral and rock samples. The procedure is based on the fusion of several aliquots of the fluorapatite standard with a well-spaced weight distribution in order to obtain a good correlation in coordinates of {sup 4}He peak height versus fluorapatite weight. The weight is then converted to moles using the accepted mineral age (31.4 Ma) and appropriate formula. Experimental peak height of {sup 4}He for the unknown samples are converted to moles with the regression determined for fluorapatite. The procedure is fast and inexpensive, and both precision and accuracy are always below 10% and usually about 3-5%.

  4. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  5. Calibration of the solar radio spectrometer

    Institute of Scientific and Technical Information of China (English)

    TAN ChengMing; YAN YiHua; TAN BaoLin; XU GuiRong

    2009-01-01

    This paper shows some improvements and new results of calibration of Chinese solar radio spectrom-eter by analyzing the daily calibration data recorded in the period of 1997-2007. First, the calibration coefficient is fitted for three bands (1.0-2.0 GHz, 2.6-3.8 GHz, 5.2-7.6 GHz) of the spectrometer by using the moving-average method confined by the property of the daily calibration data. By this calibration coefficient, the standard deviation of the calibration result was less than 10 sfu for 95% frequencies of 2.6-3.8 GHz band in 2003. This result is better than that calibrated with the constant coefficient. Second, the calibration coefficient is found in good correlation with local air temperature for most frequencies of 2.6-3.8 GHz band. Moreover, these results are helpful in the research of the quiet solar radio emission.

  6. Calibration of the solar radio spectrometer

    Institute of Scientific and Technical Information of China (English)

    2009-01-01

    This paper shows some improvements and new results of calibration of Chinese solar radio spectrometer by analyzing the daily calibration data recorded in the period of 1997-2007. First, the calibration coefficient is fitted for three bands (1.0-2.0 GHz, 2.6-3.8 GHz, 5.2-7.6 GHz) of the spectrometer by using the moving-average method confined by the property of the daily calibration data. By this calibration coefficient, the standard deviation of the calibration result was less than 10 sfu for 95% frequencies of 2.6-3.8 GHz band in 2003. This result is better than that calibrated with the constant coefficient. Second, the calibration coefficient is found in good correlation with local air temperature for most frequencies of 2.6-3.8 GHz band. Moreover, these results are helpful in the research of the quiet solar radio emission.

  7. GC-MS校准用异辛烷中八氟萘标准溶液的研制%Prepareation of Standard Solution of Octalfuorona phthalene in Isooctane for GC-MS Calibration

    Institute of Scientific and Technical Information of China (English)

    李艳玲; 刘霞; 张文申; 王学琴; 冀克俭; 赵晓刚

    2014-01-01

    采用重量-容量法研制了气相色谱-质谱(GC-MS)校准用100.1 pg/μL异辛烷中八氟萘溶液标准物质,分别用F检验和回归曲线法对研制的溶液标准物质进行了均匀性和稳定性检验,对定值结果的不确定度进行了评定。结果表明,研制的异辛烷中八氟萘溶液标准物质具有良好的均匀性和稳定性,定值结果的相对扩展不确定度为3%(k=2)。该标准物质可用于GC-MS联用仪的EI源及负CI源的信噪比校准。%100.1 pg/μL standard solution of octafluoronaphthalene in isooctane for GC-MS calibration was developed by weight-volumetric method. Uniformity and stability of standard solution were examined by F test and regression curve method respectively. The result showed that the uniformity and stability of the standard solution were good. The uncertainty of concentration value was assessed. The relative expanded uncertainty of concentration value was 3%(k=2). This CRM may be used to calibrate EI source and CI negative ion sources of GC-MS spectrometer.

  8. Calibrating Fundamental British Values: How Head Teachers Are Approaching Appraisal in the Light of the Teachers' Standards 2012, Prevent and the Counter-Terrorism and Security Act, 2015

    Science.gov (United States)

    Revell, Lynn; Bryan, Hazel

    2016-01-01

    In requiring that teachers should "not undermine fundamental British values (FBV)," a phrase originally articulated in the Home Office counter-terrorism document, Prevent, the Teachers' Standards has brought into focus the nature of teacher professionalism. Teachers in England are now required to promote FBV within and outside school,…

  9. 关于第十三批磷酸组织胺国家对照品的协作标定%Collaborative Calibration of the 1 3 th National Reference Standard for Histamine Phosphate

    Institute of Scientific and Technical Information of China (English)

    刘倩; 吴彦霖; 张媛; 贺庆; 高华; 胡宇驰; 芮菁; 祝清芬

    2014-01-01

    Obj ective To prepare and to standardize national reference standard for histamine phosphate (150510-201313).Methods The potencies of the national reference standard for histamine phosphate were assayed by the 12th national reference standard for histamine phosphate (150510-200412)with the cat blood pressure method.Results The potency of histamine phosphate was calculated as 9 9.1% based on consolidation of the calibration results from four laboratories.Conclusion The 13th national reference standard for histamine phosphate was established to replace the 12th reference standard.The determined relative potency of histamine phosphate is 100%,that is,1 mg histamine phosphate is equivalent to 0.342 mg histamine.%目的:制备、标定第十三批磷酸组织胺国家对照品(批号150510~201313)。方法用第十二批磷酸组织胺国家对照品(批号150510~200412)为标准,采用猫血压法标定。结果此批待标品经4个实验室的标定结果合并计算效价为99.1%。结论此批产品可以作为第十三批磷酸组织胺国家对照品使用。此批国家对照品相对效价定为100%,即1 mg 的磷酸组织胺相当于0.342 mg 的组织胺。

  10. Absolute sensitivity calibration of extreme ultraviolet photoresists

    Energy Technology Data Exchange (ETDEWEB)

    Jones, Juanita; Naulleau, Patrick P.; Gullikson, Eric M.; Aquila, Andrew; George, Simi; Niakoula, Dimitra

    2008-05-16

    One of the major challenges facing the commercialization of extreme ultraviolet (EUV) lithography remains simultaneously achieving resist sensitivity, line-edge roughness, and resolution requirement. Sensitivity is of particular concern owing to its direct impact on source power requirements. Most current EUV exposure tools have been calibrated against a resist standard with the actual calibration of the standard resist dating back to EUV exposures at Sandia National Laboratories in the mid 1990s. Here they report on an independent sensitivity calibration of two baseline resists from the SEMATECH Berkeley MET tool performed at the Advanced Light Source Calibrations and Standards beamline. The results show the baseline resists to be approximately 1.9 times faster than previously thought based on calibration against the long standing resist standard.

  11. Final report on supplementary comparison of the calibration of a 5 kg stainless steel standard weight between INDECOPI-PERÚ and CEM-ESPAÑA (SIM.M.M-S10)

    Science.gov (United States)

    Quiroga Rojas, Aldo; Nieves Medina, María

    2012-01-01

    This report summarizes the results of the supplementary comparison SIM.M.M-S10 of a 5 kg stainless steel mass standard between INDECOPI (Peru) and CEM (Spain).The objective of the comparison was to demonstrate the metrological equivalence between the two laboratories. The results of the comparison will be used to support the Calibration and Measurement Capabilities (CMCs) declared by the institutes at 5 kg. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the SIM, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  12. Calibration of Avent Wind IRIS SN 01030167

    DEFF Research Database (Denmark)

    Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  13. Calibration of Nacelle-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  14. Calibration of Nacelle-based Lidar instrument

    DEFF Research Database (Denmark)

    Georgieva Yankova, Ginka; Courtney, Michael

    This report presents the result of the lidar calibration performed for a four-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark.Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements...... with measurement uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  15. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  16. SNR 1E 0102.2-7219 as an X-ray calibration standard in the 0.5-1.0 keV bandpass and its application to the CCD instruments aboard Chandra, Suzaku, Swift and XMM-Newton

    Science.gov (United States)

    Plucinsky, Paul P.; Beardmore, Andrew P.; Foster, Adam; Haberl, Frank; Miller, Eric D.; Pollock, Andrew M. T.; Sembay, Steve

    2017-01-01

    Context. The flight calibration of the spectral response of charge-coupled device (CCD) instruments below 1.5 keV is difficult in general because of the lack of strong lines in the on-board calibration sources typically available. This calibration is also a function of time due to the effects of radiation damage on the CCDs and/or the accumulation of a contamination layer on the filters or CCDs. Aims: We desire a simple comparison of the absolute effective areas of the current generation of CCD instruments onboard the following observatories: Chandra ACIS-S3, XMM-Newton (EPIC-MOS and EPIC-pn), Suzaku XIS, and Swift XRT and a straightforward comparison of the time-dependent response of these instruments across their respective mission lifetimes. Methods: We have been using 1E 0102.2-7219, the brightest supernova remnant in the Small Magellanic Cloud, to evaluate and modify the response models of these instruments. 1E 0102.2-7219 has strong lines of O, Ne, and Mg below 1.5 keV and little or no Fe emission to complicate the spectrum. The spectrum of 1E 0102.2-7219 has been well-characterized using the RGS gratings instrument on XMM-Newton and the HETG gratings instrument on Chandra. As part of the activities of the International Astronomical Consortium for High Energy Calibration (IACHEC), we have developed a standard spectral model for 1E 0102.2-7219 and fit this model to the spectra extracted from the CCD instruments. The model is empirical in that it includes Gaussians for the identified lines, an absorption component in the Galaxy, another absorption component in the SMC, and two thermal continuum components with different temperatures. In our fits, the model is highly constrained in that only the normalizations of the four brightest lines/line complexes (the O vii Heα triplet, O viii Lyα line, the Ne ix Heα triplet, and the Ne x Lyα line) and an overall normalization are allowed to vary, while all other components are fixed. We adopted this approach to

  17. 用于可吸入颗粒物采样器校准的标准微粒研制和定值%Development and certified value of standard particulate matters for calibrating inhalable particulate matter sampler

    Institute of Scientific and Technical Information of China (English)

    卓春凤; 李绍海; 曹放; 黄玉虎; 杨书义

    2013-01-01

    An inhalable particulate matter (PM10) sampler is the necessary device for monitoring PM10 in air. The accuracy of the device is a primary character in evaluation of the apparatus' metro-logical performance. Spherical standard particulate matter is ideal for the calibration of samplers. Five kinds of monodisperse microsphere size particulate matters which are made from polystyrene cross-linked with divinylbenzene were developed and certified. The technical specifications of the developed standard particulate matters meet the requirements for cutting test of inhalable particulate matter PM10 sampler.%可吸入颗粒物PM10采样器(切割器)适用于大气中PM10可吸入颗粒物监测,采样器的准确性是评价其计量性能的主要指标.校准采样器需使用球形标准微粒,介绍5种微米级单分散交联聚苯乙烯球形微粒的研制和定值.所研制的5种微粒粒度标准物质,其技术指标满足我国对可吸入颗粒物PM10采样器的切割测试要求.

  18. Proposta de padronização para aferição de equipamentos de fototerapia Standardizing the calibration of phototherapy devices - a proposal

    Directory of Open Access Journals (Sweden)

    Fernando Perazzini Facchini

    2001-04-01

    Full Text Available OBJETIVO: propor método de avaliação e manutenção da eficiência dos equipamentos mais comumente usados em nosso meio, de maneira a uniformizar os resultados obtidos em tratamentos fototerápicos por diferentes serviços. MATERIAL E MÉTODOS: o radiômetro/fotômetro utilizado na medida da irradiância espectral é de fabricação nacional e atende aos requisitos básicos indispensáveis. As padronizações de medida procuram ser facilmente aplicáveis e reprodutíveis aos aparelhos de fototerapia em uso no país. O material necessário às aferições, excetuando-se o radiômetro, pode ser facilmente confeccionado em qualquer serviço de neonatologia. Procuramos nos trabalhos publicados nos últimos 10 anos no Medline os pertinentes a esta revisão. RESULTADO: propõe-se que a irradiância relatada nos trabalhos de fototerapia seja fornecida como Irradiância Espectral Média e descreve-se como obtê-la, tanto em fototerapias com fontes fluorescentes como halógenas. COMENTÁRIOS: são descritos os componentes dos radiômetros/fotômetros, cuja compreensão é importante para que possamos perceber variações na sensibilidade e diferenças significativas na irradância quando as faixas de leitura dos radiômetros são um pouco diferentes. A adoção de um tipo de equipamento padrão facilitaria muito a comparação de resultados obtidos por diferentes serviços. São ainda comentadas as características de diferentes equipamentos de fototerapia que podem interferir na mensuração da irradiância dos mesmos.OBJECTIVE: to propose a method to evaluate and maintain the efficiency of equipment commonly used in Brazil in order to obtain uniform results in phototherapy treatments in different services. MATERIAL AND METHODS: the radiometer/photometer used to measure spectral irradiance is locally manufactured and fulfills basic requirements. Measure standardization seeks to be applicable and reproducible to phototherapy devices employed in Brazil

  19. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Joseph Rovani; Mark Sanderson

    2008-02-29

    Mercury continuous emissions monitoring systems (CEMS) are being implemented in over 800 coal-fired power plant stacks. The power industry desires to conduct at least a full year of monitoring before the formal monitoring and reporting requirement begins on January 1, 2009. It is important for the industry to have available reliable, turnkey equipment from CEM vendors. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The generators are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 requires that calibration be performed with NIST-traceable standards (Federal Register 2007). Traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued an interim traceability protocol for elemental mercury generators (EPA 2007). The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The document is divided into two separate sections. The first deals with the qualification of generators by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the generator models that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma/mass spectrometry performed by NIST in Gaithersburg, MD. The

  20. Evaluation of the energy dependence of ionization chambers pencil type calibrated beam tomography standards; Avaliacao da dependencia energetica de camaras de ionizacao do tipo lapis calibradas em feixes padroes de tomografia

    Energy Technology Data Exchange (ETDEWEB)

    Fontes, Ladyjane Pereira; Potiens, Maria da Penha A., E-mail: lpfontes@ipen.br [Instituto de Pesquisas Energeticas e Nucleres (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2015-07-01

    The Instrument Calibration Laboratory of IPEN (LCI - IPEN) performs calibrations of pencil-type ionization chambers (IC) used in measures of dosimetric survey on clinical systems of Computed Tomography (CT). Many users make mistakes when using a calibrated ionization chamber in their CT dosimetry systems. In this work a methodology for determination of factors of correction for quality (Kq) through the calibration curve that is specific for each ionization chamber was established. Furthermore, it was possible to demonstrate the energy dependence on an pencil-type Ionization Chamber(IC) calibrated at the LCI - IPEN. (author)

  1. Spectral calibration for convex grating imaging spectrometer

    Science.gov (United States)

    Zhou, Jiankang; Chen, Xinhua; Ji, Yiqun; Chen, Yuheng; Shen, Weimin

    2013-12-01

    Spectral calibration of imaging spectrometer plays an important role for acquiring target accurate spectrum. There are two spectral calibration types in essence, the wavelength scanning and characteristic line sampling. Only the calibrated pixel is used for the wavelength scanning methods and he spectral response function (SRF) is constructed by the calibrated pixel itself. The different wavelength can be generated by the monochromator. The SRF is constructed by adjacent pixels of the calibrated one for the characteristic line sampling methods. And the pixels are illuminated by the narrow spectrum line and the center wavelength of the spectral line is exactly known. The calibration result comes from scanning method is precise, but it takes much time and data to deal with. The wavelength scanning method cannot be used in field or space environment. The characteristic line sampling method is simple, but the calibration precision is not easy to confirm. The standard spectroscopic lamp is used to calibrate our manufactured convex grating imaging spectrometer which has Offner concentric structure and can supply high resolution and uniform spectral signal. Gaussian fitting algorithm is used to determine the center position and the Full-Width-Half-Maximum(FWHM)of the characteristic spectrum line. The central wavelengths and FWHMs of spectral pixels are calibrated by cubic polynomial fitting. By setting a fitting error thresh hold and abandoning the maximum deviation point, an optimization calculation is achieved. The integrated calibration experiment equipment for spectral calibration is developed to enhance calibration efficiency. The spectral calibration result comes from spectral lamp method are verified by monochromator wavelength scanning calibration technique. The result shows that spectral calibration uncertainty of FWHM and center wavelength are both less than 0.08nm, or 5.2% of spectral FWHM.

  2. Comparison between external and internal standard calibration in the validation of an analytical method for 1-hydroxypyrene in human urine by high-performance liquid chromatography/tandem mass spectrometry.

    Science.gov (United States)

    Pigini, D; Cialdella, A M; Faranda, P; Tranfo, G

    2006-01-01

    1-Hydroxypyrene is a metabolite of pyrene, a member of the class of polycyclic aromatic hydrocarbons (PAHs) whose toxic properties in some cases include carcinogenicity. The determination of 1-hydroxypyrene in human urine is used as a biological indicator for exposure to PAHs, which is related to the combustion of organic materials, like smoking, living in urban environments, and eating grilled or smoked food. The determination of 1-hydroxypyrene by high-performance liquid chromatography (HPLC) with fluorescence detection has very good sensitivity but it is not highly specific: this can reduce accuracy in the quantitative determination of low levels of analyte in a complex matrix like urine. An HPLC method that uses triple quadrupole mass detection has been validated with the objective both to improve the signal-to-noise (S/N) ratio and to achieve the maximum specificity for the analyte in those urine samples that are richer in possible inteferents. The calibration range for 1-hydroxypyrene is from 0.005-0.1 microg/L in the urine of non-smoking healthy volunteers. After solid-phase extraction, samples were analyzed by HPLC/tandem mass spectrometry (MS/MS) in the multiple reaction monitoring (MRM) mode. In order to obtain reliable results quantitative analysis must be performed by means of the internal standard method (we used deuterium-labelled 1-hydroxypyrene): the method accuracy is not less than 85%. The S/N ratio at a concentration of 0.1 microg/L is about 10, and therefore this can be considered the lowest limit of quantitation. The method performance does not change if urine samples are measured using a calibration curve prepared in methanol, thus reducing the time of analysis and costs.

  3. Calibration services for medical applications of radiation

    Energy Technology Data Exchange (ETDEWEB)

    DeWerd, L.A.

    1993-12-31

    Calibration services for the medical community applications of radiation involve measuring radiation precisely and having traceability to the National Institute of Standards and Technology (NIST). Radiation therapy applications involve the use of ionization chambers and electrometers for external beams and well-type ionization chamber systems as well as radioactive sources for brachytherapy. Diagnostic x-ray applications involve ionization chamber systems and devices to measure other parameters of the x-ray machine, such as non-invasive kVp meters. Calibration laboratories have been established to provide radiation calibration services while maintaining traceability to NIST. New radiation applications of the medical community spur investigation to provide the future calibration needs.

  4. Tank calibration; Arqueacao de tanques

    Energy Technology Data Exchange (ETDEWEB)

    Chan, Ana [PETROBRAS S.A., Rio de Janeiro, RJ (Brazil)

    2003-07-01

    This work relates the analysis of the norms ISO (International Organization for Standardization) for calibration of vertical cylindrical tanks used in fiscal measurement, established on Joint Regulation no 1 of June 19, 2000 between the ANP (National Agency of Petroleum) and the INMETRO (National Institute of Metrology, Normalization and Industrial Quality). In this work a comparison between norms ISO and norms published by the API (American Petroleum Institute) and the IP (Institute of Petroleum) up to 2001 was made. It was concluded that norms ISO are wider than norms API, IP, and INMETRO methods in the calibration of vertical cylindrical tanks. (author)

  5. Calibration Monitor for Dark Energy Experiments

    Energy Technology Data Exchange (ETDEWEB)

    Kaiser, M. E.

    2009-11-23

    The goal of this program was to design, build, test, and characterize a flight qualified calibration source and monitor for a Dark Energy related experiment: ACCESS - 'Absolute Color Calibration Experiment for Standard Stars'. This calibration source, the On-board Calibration Monitor (OCM), is a key component of our ACCESS spectrophotometric calibration program. The OCM will be flown as part of the ACCESS sub-orbital rocket payload in addition to monitoring instrument sensitivity on the ground. The objective of the OCM is to minimize systematic errors associated with any potential changes in the ACCESS instrument sensitivity. Importantly, the OCM will be used to monitor instrument sensitivity immediately after astronomical observations while the instrument payload is parachuting to the ground. Through monitoring, we can detect, track, characterize, and thus correct for any changes in instrument senstivity over the proposed 5-year duration of the assembled and calibrated instrument.

  6. SNR 1E 0102.2-7219 as an X-ray Calibration Standard in the 0.5-1.0 keV Bandpass and Its Application to the CCD Instruments aboard Chandra, Suzaku, Swift and XMM-Newton

    CERN Document Server

    Plucinsky, Paul P; Foster, Adam; Haberl, Frank; Miller, Eric D; Pollock, A M T; Sembay, Steve

    2016-01-01

    We desire a simple comparison of the absolute effective areas of the current generation of CCD instruments onboard the following observatories: Chandra ACIS-S3, XMM-Newton (EPIC-MOS and EPIC-pn), Suzaku XIS, and Swift XRT and a straightforward comparison of the time-dependent response of these instruments across their respective mission lifetimes. We have been using 1E 0102.2-7219, the brightest supernova remnant in the Small Magellanic Cloud, to evaluate and modify the response models of these instruments. 1E 0102.2-7219 has strong lines of O, Ne, and Mg below 1.5 keV and little or no Fe emission to complicate the spectrum. As part of the activities of the International Astronomical Consortium for High Energy Calibration (IACHEC), we have developed a standard spectral model for 1E 0102.2-7219. The model is empirical in that it includes Gaussians for the identified lines, an absorption component in the Galaxy, another absorption component in the SMC, and two thermal continuum components. In our fits, the mode...

  7. The CIEMAT Neutron Standards Laboratory (LPN): National Reference Facility for the Calibration of Neutron Measurement Equipment; El Laboratorio de Patrones Neutronicos del CIEMAT: instalacion de referencia nacional para la calibracion de equipos de medida neutronica

    Energy Technology Data Exchange (ETDEWEB)

    Mendez, R.

    2014-04-01

    This is a new facility that, in keeping with the recommendations of standard ISO 8529:1, has 20m5-Be and 252Cf neutron sources stored in water inside a 9 m x 7.5 m x 8 m Irradiation Room with 1.25 m concrete walls that is accessed through a 6.5 ton sliding shielded door. These sources are remotely handled from the Control Room by means of a system that places them in the geometric center of the room and aligns them with the equipment or materials to be irradiated arranged on a base plate. The facility has a series of interlocks that guarantee safety during operations with sources, and the verification measurements have demonstrated the suitability of the shielding used. The LPN will be the answer to the demand for a national laboratory for neutron measurement equipment calibration and equipment and material irradiation under controlled conditions, and it will position our country at the same level as our European partners. (Author)

  8. Trinocular Calibration Method Based on Binocular Calibration

    Directory of Open Access Journals (Sweden)

    CAO Dan-Dan

    2012-10-01

    Full Text Available In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error of the global calibration of the two camera pairs in the experiments can be as low as 0.277% and 0.328% respectively. Experiment results show that this method is feasible, simple and effective, and has high precision.

  9. 40 CFR 1065.790 - Mass standards.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 32 2010-07-01 2010-07-01 false Mass standards. 1065.790 Section 1065... ENGINE-TESTING PROCEDURES Engine Fluids, Test Fuels, Analytical Gases and Other Calibration Standards § 1065.790 Mass standards. (a) PM balance calibration weights. Use PM balance calibration weights...

  10. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    Science.gov (United States)

    Cornic, Philippe; Illoul, Cédric; Cheminet, Adam; Le Besnerais, Guy; Champagnat, Frédéric; Le Sant, Yves; Leclaire, Benjamin

    2016-09-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data.

  11. Sensor Calibration Design Based on D-Optimality Criterion

    Directory of Open Access Journals (Sweden)

    Hajiyev Chingiz

    2016-09-01

    Full Text Available In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM. As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.

  12. Cumulative sum quality control for calibrated breast density measurements

    Energy Technology Data Exchange (ETDEWEB)

    Heine, John J.; Cao Ke; Beam, Craig [Cancer Prevention and Control Division, Moffitt Cancer Center, 12902 Magnolia Drive, Tampa, Florida 33612 (United States); Division of Epidemiology and Biostatistics, School of Public Health, University of Illinois at Chicago, 1603 W. Taylor St., Chicago, Illinois 60612 (United States)

    2009-12-15

    Purpose: Breast density is a significant breast cancer risk factor. Although various methods are used to estimate breast density, there is no standard measurement for this important factor. The authors are developing a breast density standardization method for use in full field digital mammography (FFDM). The approach calibrates for interpatient acquisition technique differences. The calibration produces a normalized breast density pixel value scale. The method relies on first generating a baseline (BL) calibration dataset, which required extensive phantom imaging. Standardizing prospective mammograms with calibration data generated in the past could introduce unanticipated error in the standardized output if the calibration dataset is no longer valid. Methods: Sample points from the BL calibration dataset were imaged approximately biweekly over an extended timeframe. These serial samples were used to evaluate the BL dataset reproducibility and quantify the serial calibration accuracy. The cumulative sum (Cusum) quality control method was used to evaluate the serial sampling. Results: There is considerable drift in the serial sample points from the BL calibration dataset that is x-ray beam dependent. Systematic deviation from the BL dataset caused significant calibration errors. This system drift was not captured with routine system quality control measures. Cusum analysis indicated that the drift is a sign of system wear and eventual x-ray tube failure. Conclusions: The BL calibration dataset must be monitored and periodically updated, when necessary, to account for sustained system variations to maintain the calibration accuracy.

  13. The calibration of PIXIE

    Science.gov (United States)

    Fixsen, D. J.; Chuss, D. T.; Kogut, Alan; Mirel, Paul; Wollack, E. J.

    2016-07-01

    The FIRAS instrument demonstrated the use of an external calibrator to compare the sky to an instrumented blackbody. The PIXIE calibrator is improved from -35 dB to -65 dB. Another significant improvement is the ability to insert the calibrator into either input of the FTS. This allows detection and correction of additional errors, reduces the effective calibration noise by a factor of 2, eliminates an entire class of systematics and allows continuous observations. This paper presents the design and use of the PIXIE calibrator.

  14. Calibration of Nanopositioning Stages

    Directory of Open Access Journals (Sweden)

    Ning Tan

    2015-12-01

    Full Text Available Accuracy is one of the most important criteria for the performance evaluation of micro- and nanorobots or systems. Nanopositioning stages are used to achieve the high positioning resolution and accuracy for a wide and growing scope of applications. However, their positioning accuracy and repeatability are not well known and difficult to guarantee, which induces many drawbacks for many applications. For example, in the mechanical characterisation of biological samples, it is difficult to perform several cycles in a repeatable way so as not to induce negative influences on the study. It also prevents one from controlling accurately a tool with respect to a sample without adding additional sensors for closed loop control. This paper aims at quantifying the positioning repeatability and accuracy based on the ISO 9283:1998 standard, and analyzing factors influencing positioning accuracy onto a case study of 1-DoF (Degree-of-Freedom nanopositioning stage. The influence of thermal drift is notably quantified. Performances improvement of the nanopositioning stage are then investigated through robot calibration (i.e., open-loop approach. Two models (static and adaptive models are proposed to compensate for both geometric errors and thermal drift. Validation experiments are conducted over a long period (several days showing that the accuracy of the stage is improved from typical micrometer range to 400 nm using the static model and even down to 100 nm using the adaptive model. In addition, we extend the 1-DoF calibration to multi-DoF with a case study of a 2-DoF nanopositioning robot. Results demonstrate that the model efficiently improved the 2D accuracy from 1400 nm to 200 nm.

  15. Electromagnetic Calorimeter Calibration with $\\pi^{0}$

    CERN Multimedia

    Puig Navarro, A

    2009-01-01

    Several methods can be used in order to achieve precise calibration of the LHCb Electromagnetic Calorimeter (ECAL) once reasonable cell equalization has been reached. At low transverse energy, the standard calibration procedure is an iterative method based on the fit of the $\\gamma\\gamma$ invariant mass distribution for each cell of the decay $\\pi^{0}\\to\\gamma\\gamma$ with resolved photons. A new technique for generating the combinatorial background of such decays directly from data has been developed. Knowledge of the background could allow an alternative calibration method based on a event by event fit of the same $\\gamma\\gamma$ invariant mass distribution where contributions from groups of cells are considered in a single fit. The background generation procedure and this possible new calibration method are presented in this poster, in addition to an overview of the LHCb Calorimetry system and ECAL calibration techniques.

  16. Simple transfer calibration method for a Cimel Sun-Moon photometer: calculating lunar calibration coefficients from Sun calibration constants.

    Science.gov (United States)

    Li, Zhengqiang; Li, Kaitao; Li, Donghui; Yang, Jiuchun; Xu, Hua; Goloub, Philippe; Victori, Stephane

    2016-09-20

    The Cimel new technologies allow both daytime and nighttime aerosol optical depth (AOD) measurements. Although the daytime AOD calibration protocols are well established, accurate and simple nighttime calibration is still a challenging task. Standard lunar-Langley and intercomparison calibration methods both require specific conditions in terms of atmospheric stability and site condition. Additionally, the lunar irradiance model also has some known limits on its uncertainty. This paper presents a simple calibration method that transfers the direct-Sun calibration constant, V0,Sun, to the lunar irradiance calibration coefficient, CMoon. Our approach is a pure calculation method, independent of site limits, e.g., Moon phase. The method is also not affected by the lunar irradiance model limitations, which is the largest error source of traditional calibration methods. Besides, this new transfer calibration approach is easy to use in the field since CMoon can be obtained directly once V0,Sun is known. Error analysis suggests that the average uncertainty of CMoon over the 440-1640 nm bands obtained with the transfer method is 2.4%-2.8%, depending on the V0,Sun approach (Langley or intercomparison), which is comparable with that of lunar-Langley approach, theoretically. In this paper, the Sun-Moon transfer and the Langley methods are compared based on site measurements in Beijing, and the day-night measurement continuity and performance are analyzed.

  17. HPS instrument calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Masse, F.X; Eisenhower, E.H.; Swinth, K.L.

    1993-12-31

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory.

  18. Calibration of routine dosimeters in radiation processing: Validation procedure for in-plant calibration

    Directory of Open Access Journals (Sweden)

    Šećerov Bojana Lj.

    2011-01-01

    Full Text Available The essential prerequisite of radiation dosimetry is to provide quality assurance and documentation that the irradiation procedure has been carried out according to the specification requirement of correct calibration of the chosen dosimetry system. At the Radiation Plant of the Vinča Institute of Nuclear Sciences we compared two recommended protocols of irradiation procedures in the calibration of dosimetry systems in radiation processing: (1 by irradiation of routine dosimeters (ethanol-chlorobenzene - ECB at the calibration laboratory and (2, by in-plant calibration with alanine transfer - dosimeters. The critical point for in-plant calibration is irradiation geometry, so we carefully positioned the phantom carrying both dosimeters in order to minimize dose gradients across the sample. The analysis of results obtained showed that the difference among determined absorbed doses for the construction of calibration curves between these two methods, (alanine vs. ECB, is less than 1%. The difference in combined standard uncertainty for each calibration procedure is 0.1%. These results demonstrate that our in-plant calibration is as good as calibration by irradiation at the calibration laboratory and validates our placement of the irradiation phantom during irradiation.

  19. Preliminary evaluation of a Neutron Calibration Laboratory

    Energy Technology Data Exchange (ETDEWEB)

    Alvarenga, Talysson S.; Neves, Lucio P.; Perini, Ana P.; Sanches, Matias P.; Mitake, Malvina B.; Caldas, Linda V.E., E-mail: talvarenga@ipen.br, E-mail: lpneves@ipen.br, E-mail: aperini@ipen.br, E-mail: msanches@ipen.br, E-mail: mbmitake@ipen.br, E-mail: lcaldas@ipen.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Federico, Claudio A., E-mail: claudiofederico@ieav.cta.br [Instituto de Estudos Avancados (IEAv/DCTA), Sao Jose dos Campos, SP (Brazil). Dept. de Ciencia e Tecnologia Aeroespacial

    2013-07-01

    In the past few years, Brazil and several other countries in Latin America have experimented a great demand for the calibration of neutron detectors, mainly due to the increase in oil prospection and extraction. The only laboratory for calibration of neutron detectors in Brazil is localized at the Institute for Radioprotection and Dosimetry (IRD/CNEN), Rio de Janeiro, which is part of the IAEA SSDL network. This laboratory is the national standard laboratory in Brazil. With the increase in the demand for the calibration of neutron detectors, there is a need for another calibration services. In this context, the Calibration Laboratory of IPEN/CNEN, Sao Paulo, which already offers calibration services of radiation detectors with standard X, gamma, beta and alpha beams, has recently projected a new calibration laboratory for neutron detectors. In this work, the ambient equivalent dose rate (H⁎(10)) was evaluated in several positions inside and around this laboratory, using Monte Carlo simulation (MCNP5 code), in order to verify the adequateness of the shielding. The obtained results showed that the shielding is effective, and that this is a low-cost methodology to improve the safety of the workers and evaluate the total staff workload. (author)

  20. Calibration of Ground-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  1. Calibration of Ground -based Lidar instrument

    DEFF Research Database (Denmark)

    Villanueva, Héctor; Yordanova, Ginka

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  2. Use of Ethanol/Diesel Blend and Advanced Calibration Methods to Satisfy Euro 5 Emission Standards without DPF Utilisation d’un carburant Diesel éthanolé à l’aide de méthodes de calibration avancées afin de satisfaire les normes Euro 5 sans filtre à particules

    Directory of Open Access Journals (Sweden)

    Magand S.

    2011-11-01

    innovative calibration methods, based on the simultaneous optimisation of engine basic settings and cold correction maps, are introduced in order to better suit to the new formulation impact on combustion and catalyst light-off and to drop off engine-out unburned hydrocarbons and carbon monoxide emissions. This stage allows pushing forward the work on test bed facilities in order to reduce the amount of vehicle tests. Tests on a chassis dynamometer are only used to validate the engine test bed results and to perform final tuning of cold correction maps. This alternative blend shows potential to achieve Euro 5 standard with Euro 4 Diesel vehicle configuration, without any hardware modification and without a Diesel particulate filter in the exhaust line. Such an innovative fuel formulation seems to be an interesting answer to the trade-off in the forthcoming years between cost and emissions reduction to achieve sustainable mobility. The presented calibration methods and tools allow to fully take advantage of this alternative fuel in a reduced time scale. L’utilisation des biocarburants s’est développée durant ces dernières années de façon importante afin de diversifier les sources d’énergies et de limiter la hausse des émissions de gaz à effet de serre du secteur des transports. L’un des carburants renouvelables les plus adaptés à une production de masse est l’éthanol. Celui-ci est aujourd’hui principalement utilisé dans les moteurs à allumage commandé, alors que la part des véhicules Diesel sur le marché européen est de l’ordre de 60 %. Ce constat nous a incité à proposer une formulation innovante utilisant de l’éthanol pour les applications Diesel. Les principaux verrous technologiques pour cette utilisation sont la miscibilité, la température d’éclair, la lubrification ou encore l’indice de cétane. Des travaux ont été réalisés pour optimiser la formulation contenant de l’éthanol, des biodiesels de première et seconde g

  3. PACS photometer calibration block analysis

    CERN Document Server

    Moór, A; Kiss, Cs; Balog, Z; Billot, N; Marton, G

    2013-01-01

    The absolute stability of the PACS bolometer response over the entire mission lifetime without applying any corrections is about 0.5% (standard deviation) or about 8% peak-to-peak. This fantastic stability allows us to calibrate all scientific measurements by a fixed and time-independent response file, without using any information from the PACS internal calibration sources. However, the analysis of calibration block observations revealed clear correlations of the internal source signals with the evaporator temperature and a signal drift during the first half hour after the cooler recycling. These effects are small, but can be seen in repeated measurements of standard stars. From our analysis we established corrections for both effects which push the stability of the PACS bolometer response to about 0.2% (stdev) or 2% in the blue, 3% in the green and 5% in the red channel (peak-to-peak). After both corrections we still see a correlation of the signals with PACS FPU temperatures, possibly caused by parasitic h...

  4. Distributed Radio Interferometric Calibration

    CERN Document Server

    Yatawatta, Sarod

    2015-01-01

    Increasing data volumes delivered by a new generation of radio interferometers require computationally efficient and robust calibration algorithms. In this paper, we propose distributed calibration as a way of improving both computational cost as well as robustness in calibration. We exploit the data parallelism across frequency that is inherent in radio astronomical observations that are recorded as multiple channels at different frequencies. Moreover, we also exploit the smoothness of the variation of calibration parameters across frequency. Data parallelism enables us to distribute the computing load across a network of compute agents. Smoothness in frequency enables us reformulate calibration as a consensus optimization problem. With this formulation, we enable flow of information between compute agents calibrating data at different frequencies, without actually passing the data, and thereby improving robustness. We present simulation results to show the feasibility as well as the advantages of distribute...

  5. OLI Radiometric Calibration

    Science.gov (United States)

    Markham, Brian; Morfitt, Ron; Kvaran, Geir; Biggar, Stuart; Leisso, Nathan; Czapla-Myers, Jeff

    2011-01-01

    Goals: (1) Present an overview of the pre-launch radiance, reflectance & uniformity calibration of the Operational Land Imager (OLI) (1a) Transfer to orbit/heliostat (1b) Linearity (2) Discuss on-orbit plans for radiance, reflectance and uniformity calibration of the OLI

  6. WFPC2 Polarization Calibration

    Science.gov (United States)

    Biretta, J.; McMaster, M.

    1997-12-01

    We derive a detailed calibration for WFPC2 polarization data which is accurate to about 1.5%. We begin by computing polarizer flats, and show how they are applied to data. A physical model for the polarization effects of the WFPC2 optics is then created using Mueller matricies. This model includes corrections for the instrumental polarization (diattenuation and phase retardance) of the pick-off mirror, as well as the high cross-polarization transmission of the polarizer filter. We compare this model against the on-orbit observations of polarization calibrators, and show it predicts relative counts in the different polarizer/aperture settings to 1.5% RMS accuracy. We then show how this model can be used to calibrate GO data, and present two WWW tools which allow observers to easily calibrate their data. Detailed examples are given illustrationg the calibration and display of WFPC2 polarization data. In closing we describe future plans and possible improvements.

  7. Dry calibration of ultrasonic gas flow meters

    Energy Technology Data Exchange (ETDEWEB)

    De Boer, G.; Lansing, J.

    1997-07-01

    At present in most European countries it is customary that turbine meters, or the newer ultrasonic gas flow meters, when used in fiscal metering or custody transfer metering applications, are calibrated in a test facility by comparison to standards or reference devices. For reason of practical and operational drawbacks, costs involved and availability of only a limited number of calibration facilities, another way of meter verification is advantageous. For orifice metering the practice of dry calibration is well established; that is, meter verification is based upon examination of the geometry and installation of the orifice plate and a function check of the read out devices. Although for turbine meters a flow (wet) calibration may be a necessity, it will be shown that ultrasonic gas flow meters can be dry calibrated in the same way as orifice meters. As a basis for the acceptance of a dry calibration procedure for ultrasonic gas flow meters, a sensitivity analysis of the relevant variables with respect to the meter's accuracy is presented. Further test results are presented that demonstrate the feasibility of the concept of dry calibration applied to ultrasonic gas flow meters. (author)

  8. Segment Based Camera Calibration

    Institute of Scientific and Technical Information of China (English)

    马颂德; 魏国庆; 等

    1993-01-01

    The basic idea of calibrating a camera system in previous approaches is to determine camera parmeters by using a set of known 3D points as calibration reference.In this paper,we present a method of camera calibration in whih camera parameters are determined by a set of 3D lines.A set of constraints is derived on camea parameters in terms of perspective line mapping.Form these constraints,the same perspective transformation matrix as that for point mapping can be computed linearly.The minimum number of calibration lines is 6.This result generalizes that of Liu,Huang and Faugeras[12] for camera location determination in which at least 8 line correspondences are required for linear computation of camera location.Since line segments in an image can be located easily and more accurately than points,the use of lines as calibration reference tends to ease the computation in inage preprocessing and to improve calibration accuracy.Experimental results on the calibration along with stereo reconstruction are reported.

  9. 对JB 4730标准中无损检测仪器设备校准或核查要求的修改建议%The Modification Suggestions for the Requirements of Calibration or Check of Non-Destructive Testing Instruments in JB 4730 Standard

    Institute of Scientific and Technical Information of China (English)

    谢常欢

    2012-01-01

    This article compares the requirements of calibration or check for nondestructive testing instruments with EN-04/15,HOKLAS Supplementary etc.,and gives the modification suggestions for the requirements of calibration or check of non-destructive testing instruments for new JB 4730 standard.%借鉴欧盟EN-04/15、香港HOKLAS增补准则第15等标准中对无损检测仪器设备校准或核查的要求,对今后新版JB 4730标准中有关无损检测仪器设备校准或核查的要求提出了修改建议。

  10. A Novel Calibrator for Electronic Transformers Based on IEC 61850

    Directory of Open Access Journals (Sweden)

    Baoxiang PAN

    2013-01-01

    Full Text Available It is necessary for electronic transformer to make calibration before putting it into practice. To solve the problems in actual calibration process, a novel electronic transformer calibrator is designed. In principle, this system adopts both the direct method and the difference method, which are two popular methods for electronic transformer calibration, by this way the application of the system is extended with its reliability improved. In the system design, based on virtual instrument technology, LabVIEW and WinPCap toolkit are used to develop the application software, and it is able to calibrate those electronic transformers following the standard of IEC 61850. In the calculation of ratio and phase error based on fast Fourier transform, a new window function is introduced, and thus the accuracy of calibration, influenced by the frequency vibration, is improved. This research provides theoretic support and practical reference to the development of intelligent calibrator for electronic transformers.

  11. Calibration Fixture For Anemometer Probes

    Science.gov (United States)

    Lewis, Charles R.; Nagel, Robert T.

    1993-01-01

    Fixture facilitates calibration of three-dimensional sideflow thermal anemometer probes. With fixture, probe oriented at number of angles throughout its design range. Readings calibrated as function of orientation in airflow. Calibration repeatable and verifiable.

  12. Calibration of Avent Wind IRIS SN. WI01030176

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  13. Air Data Calibration Facility

    Data.gov (United States)

    Federal Laboratory Consortium — This facility is for low altitude subsonic altimeter system calibrations of air vehicles. Mission is a direct support of the AFFTC mission. Postflight data merge is...

  14. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina; Webb, Craig

    2016-05-02

    This presentation provides a high-level overview of the progress on the Broadband Outdoor Radiometer Calibrations for all shortwave and longwave radiometers that are deployed by the Atmospheric Radiation Measurement program.

  15. SRHA calibration curve

    Data.gov (United States)

    U.S. Environmental Protection Agency — an UV calibration curve for SRHA quantitation This dataset is associated with the following publication: Chang, X., and D. Bouchard. Surfactant-Wrapped Multiwalled...

  16. Calibrating nacelle lidars

    DEFF Research Database (Denmark)

    Courtney, Michael

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report...... presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated...... a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam...

  17. Calibrating nacelle lidars

    OpenAIRE

    Courtney, Michael

    2013-01-01

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail.The first of these is a line of sight...

  18. TWSTFT Link Calibration Report

    Science.gov (United States)

    2015-09-01

    box calibrator with unknown but constant total delay during a calibration tour Total Delay: The total electrical delay from the antenna phase center...to the UTCp including all the devices/cables that the satellite and clock signals pass through. It numerically equals the sum of all the sub-delays...PTB. To average out the dimnal effects and measurement noise , 5-7 days of continuous measurements is required. 3 Setups at the Lab(k) The setup

  19. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  20. Energy calibration via correlation

    CERN Document Server

    Maier, Daniel

    2015-01-01

    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241 Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be le...

  1. A new irradiated quartz for beta source calibration

    DEFF Research Database (Denmark)

    Hansen, Vicki; Murray, Andrew Sean; Buylaert, Jan-Pieter;

    2015-01-01

    laboratories have used the various different batches of Risø calibration quartz for the calibration of beta and X-ray sources, but these have been largely undescribed. Here we describe in detail the preparation and luminescence characteristics of a new quartz standard, based on a North Sea beach sand collected...

  2. Automated intraoperative calibration for prostate cancer brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Kuiran Chen, Thomas; Heffter, Tamas; Lasso, Andras; Pinter, Csaba; Abolmaesumi, Purang; Burdette, E. Clif; Fichtinger, Gabor [Queen' s University, Kingston, Ontario K7L 3N6 (Canada); University of British Columbia, Vancouver, British Columbia V6T 1Z4 (Canada); Acoustic MedSystems, Inc., Champaign, Illinois 61820-3979 (United States); Queen' s University, Kingston, Ontario K7L 3N6 (Canada) and Johns Hopkins University, Baltimore, Maryland 21218-2682 (United States)

    2011-11-15

    Purpose: Prostate cancer brachytherapy relies on an accurate spatial registration between the implant needles and the TRUS image, called ''calibration''. The authors propose a new device and a fast, automatic method to calibrate the brachytherapy system in the operating room, with instant error feedback. Methods: A device was CAD-designed and precision-engineered, which mechanically couples a calibration phantom with an exact replica of the standard brachytherapy template. From real-time TRUS images acquired from the calibration device and processed by the calibration system, the coordinate transformation between the brachytherapy template and the TRUS images was computed automatically. The system instantly generated a report of the target reconstruction accuracy based on the current calibration outcome. Results: Four types of validation tests were conducted. First, 50 independent, real-time calibration trials yielded an average of 0.57 {+-} 0.13 mm line reconstruction error (LRE) relative to ground truth. Second, the averaged LRE was 0.37 {+-} 0.25 mm relative to ground truth in tests with six different commercial TRUS scanners operating at similar imaging settings. Furthermore, testing with five different commercial stepper systems yielded an average of 0.29 {+-} 0.16 mm LRE relative to ground truth. Finally, the system achieved an average of 0.56 {+-} 0.27 mm target registration error (TRE) relative to ground truth in needle insertion tests through the template in a water tank. Conclusions: The proposed automatic, intraoperative calibration system for prostate cancer brachytherapy has achieved high accuracy, precision, and robustness.

  3. A Method to Test Model Calibration Techniques

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, Ron; Polly, Ben; Neymark, Joel

    2016-08-26

    This paper describes a method for testing model calibration techniques. Calibration is commonly used in conjunction with energy retrofit audit models. An audit is conducted to gather information about the building needed to assemble an input file for a building energy modeling tool. A calibration technique is used to reconcile model predictions with utility data, and then the 'calibrated model' is used to predict energy savings from a variety of retrofit measures and combinations thereof. Current standards and guidelines such as BPI-2400 and ASHRAE-14 set criteria for 'goodness of fit' and assume that if the criteria are met, then the calibration technique is acceptable. While it is logical to use the actual performance data of the building to tune the model, it is not certain that a good fit will result in a model that better predicts post-retrofit energy savings. Therefore, the basic idea here is that the simulation program (intended for use with the calibration technique) is used to generate surrogate utility bill data and retrofit energy savings data against which the calibration technique can be tested. This provides three figures of merit for testing a calibration technique, 1) accuracy of the post-retrofit energy savings prediction, 2) closure on the 'true' input parameter values, and 3) goodness of fit to the utility bill data. The paper will also discuss the pros and cons of using this synthetic surrogate data approach versus trying to use real data sets of actual buildings.

  4. [Research on calibration transfer of NIR filter spectrophotometer].

    Science.gov (United States)

    Chen, Jia-wei; Zhou, Chang-le; Zhang, Ye-hui; Xu, Xiao-jie; Lin, Kun-hui; Ye, Nan

    2008-10-01

    Calibration transfer is an important issue to building up universal and comparable performance of spectrometer data in near infrared spectral analysis technology. Methods of slope/bias correction, direct standardization (DS), and target factor analysis (TFA) were used for the calibration transfer among five NIR filter spectrophotometers using maize as the samples. The effects of three calibration transfer methods were compared. The DS method has the best performance. The average calibration transfer difference of DS is 7.01%. This study also relates to the dependence of calibration transfer on the number of standardization samples. It was proven by experiment that the results of calibration transfer will be better as the number of samples is increased and will be generally stable when there are twenty standardization samples. However, the effect of calibration transfer attained by DS is degraded sharply when the number of standardization samples decreases to be below twenty. Moreover, slope/bias and TFA are not sensitive to the number of standardization samples.

  5. CTEPP STANDARD OPERATING PROCEDURE FOR PREPARATION OF SURROGATE RECOVERY STANDARD AND INTERNAL STANDARD SOLUTIONS FOR NEUTRAL TARGET ANALYTES (SOP-5.25)

    Science.gov (United States)

    This standard operating procedure describes the method used for preparing internal standard, surrogate recovery standard and calibration standard solutions for neutral analytes used for gas chromatography/mass spectrometry analysis.

  6. Calibration of burnup monitor installed in Rokkasho Reprocessing Plant

    Energy Technology Data Exchange (ETDEWEB)

    Oeda, Kaoru; Naito, Hirofumi; Hirota, Masanari [Japan Nuclear Fuel Co. Ltd., Rokkasho, Aomori (Japan); Natsume, Koichiro [Isogo Engineering Center, Toshiba Corporation, Yokohama, Kanagawa (Japan); Kumanomido, Hironori [Nuclear Engineering Laboratory, Toshiba Corporation, Kawasaki, Kanagawa (Japan)

    2000-06-01

    Rokkasho Reprocessing Plant uses burnup credit for criticality control at the Spent Fuel Storage Facility (SFSF) and the Dissolution Facility. A burnup monitor measures nondestructively burnup value of a spent fuel assembly and guarantees the credit for burnup. For practical reasons, a standard radiation source is not used in calibration of the burnup monitor, but the burnup values of many spent fuel assemblies are measured based on operator-declared burnup values. This paper describes the concept of burnup credit, the burnup monitor, and the calibration method. It is concluded, from the results of calibration tests, that the calibration method is valid. (author)

  7. Manual Calibration System for Daya Bay Reactor Neutrino Experiment

    Institute of Scientific and Technical Information of China (English)

    HUANG; Han-xiong; RUAN; Xi-chao; REN; Jie; LV; Yin-long; FAN; Cheng-jun; CHEN; Yan-nan; WANG; Zhao-hui; ZHOU; Zu-ying; HOU; Long; ZHANG; Jia-wen; ZHANG; Yin-hong; YU; Chao-ju; HE; Wei; ZHOU; Bin

    2012-01-01

    <正>The neutrino mixing angle θ13 with a significance of 7.7 standard deviations has been published by the Daya Bay anti-neutrino experiment collaboration in 2012. To understand the non-uniformity and the energy non-linearity of the anti-neutrino detector (AD), a calibration campaign for the AD1 with the Manual Calibration System (MCS) has been finished. The aim of this calibration plan is to deploy the calibration sources to any positions inside the Inner Acrylic Vessel (IAV), to study detail properties of AD.

  8. Calibration Under Uncertainty.

    Energy Technology Data Exchange (ETDEWEB)

    Swiler, Laura Painton; Trucano, Timothy Guy

    2005-03-01

    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  9. HAWC Timing Calibration

    CERN Document Server

    Huentemeyer, Petra; Dingus, Brenda

    2009-01-01

    The High-Altitude Water Cherenkov (HAWC) Experiment is a second-generation highsensitivity gamma-ray and cosmic-ray detector that builds on the experience and technology of the Milagro observatory. Like Milagro, HAWC utilizes the water Cherenkov technique to measure extensive air showers. Instead of a pond filled with water (as in Milagro) an array of closely packed water tanks is used. The event direction will be reconstructed using the times when the PMTs in each tank are triggered. Therefore, the timing calibration will be crucial for reaching an angular resolution as low as 0.25 degrees.We propose to use a laser calibration system, patterned after the calibration system in Milagro. Like Milagro, the HAWC optical calibration system will use ~1 ns laser light pulses. Unlike Milagro, the PMTs are optically isolated and require their own optical fiber calibration. For HAWC the laser light pulses will be directed through a series of optical fan-outs and fibers to illuminate the PMTs in approximately one half o...

  10. Photometric calibrations for 21st century science

    Energy Technology Data Exchange (ETDEWEB)

    Kent, Stephen; /Fermilab; Kaiser, Mary Elizabeth; /Johns Hopkins U.; Deustua, Susana E.; /Baltimore, Space Telescope Sci.; Smith, J.Allyn; /Austin Peay State U.; Adelman, Saul; /Citadel Military Coll.; Allam, Sahar S.; /Fermilab; Baptista, Brian; /Indiana U.; Bohlin, Ralph C.; /Baltimore, Space Telescope Sci.; Clem, James L.; /Louisiana State U.; Conley, Alex; /Colorado U.; Edelstein, Jerry; /UC, Berkeley, Space Sci. Dept. /NOAO, Tucson

    2009-02-01

    The answers to fundamental science questions in astrophysics, ranging from the history of the expansion of the universe to the sizes of nearby stars, hinge on our ability to make precise measurements of diverse astronomical objects. As our knowledge of the underlying physics of objects improves along with advances in detectors and instrumentation, the limits on our capability to extract science from measurements is set, not by our lack of understanding of the nature of these objects, but rather by the most mundane of all issues: the precision with which we can calibrate observations in physical units. In principle, photometric calibration is a solved problem - laboratory reference standards such as blackbody furnaces achieve precisions well in excess of those needed for astrophysics. In practice, however, transferring the calibration from these laboratory standards to astronomical objects of interest is far from trivial - the transfer must reach outside the atmosphere, extend over 4{pi} steradians of sky, cover a wide range of wavelengths, and span an enormous dynamic range in intensity. Virtually all spectrophotometric observations today are calibrated against one or more stellar reference sources, such as Vega, which are themselves tied back to laboratory standards in a variety of ways. This system's accuracy is not uniform. Selected regions of the electromagnetic spectrum are calibrated extremely well, but discontinuities of a few percent still exist, e.g., between the optical and infrared. Independently, model stellar atmospheres are used to calibrate the spectra of selected white dwarf stars, e.g. the HST system, but the ultimate accuracy of this system should be verified against laboratory sources. Our traditional standard star systems, while sufficient until now, need to be improved and extended in order to serve future astrophysics experiments. This white paper calls for a program to improve upon and expand the current networks of

  11. Measurement quality assurance for beta particle calibrations at NIST

    Energy Technology Data Exchange (ETDEWEB)

    Soares, C.G.; Pruitt, J.S. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    Standardized beta-particle fields have been established in an international standard and have been adopted for use in several U.S. dosimeter and instrument testing standards. Calibration methods and measurement quality assurance procedures employed at the National Institute of Standards and Technology (NIST) for beta-particle calibrations in these reference fields are discussed. The calibration facility including the NIST-automated extrapolation ionization chamber is described, and some sample results of calibrations are shown. Methods for establishing and maintaining traceability to NIST of secondary laboratories are discussed. Currently, there are problems in finding a good method for routine testing of traceability to NIST. Some examples of past testing methods are given and solutions to this problem are proposed.

  12. Calibration Systems Final Report

    Energy Technology Data Exchange (ETDEWEB)

    Myers, Tanya L.; Broocks, Bryan T.; Phillips, Mark C.

    2006-02-01

    The Calibration Systems project at Pacific Northwest National Laboratory (PNNL) is aimed towards developing and demonstrating compact Quantum Cascade (QC) laser-based calibration systems for infrared imaging systems. These on-board systems will improve the calibration technology for passive sensors, which enable stand-off detection for the proliferation or use of weapons of mass destruction, by replacing on-board blackbodies with QC laser-based systems. This alternative technology can minimize the impact on instrument size and weight while improving the quality of instruments for a variety of missions. The potential of replacing flight blackbodies is made feasible by the high output, stability, and repeatability of the QC laser spectral radiance.

  13. Ibis ground calibration

    Energy Technology Data Exchange (ETDEWEB)

    Bird, A.J.; Barlow, E.J.; Tikkanen, T. [Southampton Univ., School of Physics and Astronomy (United Kingdom); Bazzano, A.; Del Santo, M.; Ubertini, P. [Istituto di Astrofisica Spaziale e Fisica Cosmica - IASF/CNR, Roma (Italy); Blondel, C.; Laurent, P.; Lebrun, F. [CEA Saclay - Sap, 91 - Gif sur Yvette (France); Di Cocco, G.; Malaguti, E. [Istituto di Astrofisica Spaziale e Fisica-Bologna - IASF/CNR (Italy); Gabriele, M.; La Rosa, G.; Segreto, A. [Istituto di Astrofisica Spaziale e Fisica- IASF/CNR, Palermo (Italy); Quadrini, E. [Istituto di Astrofisica Spaziale e Fisica-Cosmica, EASF/CNR, Milano (Italy); Volkmer, R. [Institut fur Astronomie und Astrophysik, Tubingen (Germany)

    2003-11-01

    We present an overview of results obtained from IBIS ground calibrations. The spectral and spatial characteristics of the detector planes and surrounding passive materials have been determined through a series of calibration campaigns. Measurements of pixel gain, energy resolution, detection uniformity, efficiency and imaging capability are presented. The key results obtained from the ground calibration have been: - optimization of the instrument tunable parameters, - determination of energy linearity for all detection modes, - determination of energy resolution as a function of energy through the range 20 keV - 3 MeV, - demonstration of imaging capability in each mode, - measurement of intrinsic detector non-uniformity and understanding of the effects of passive materials surrounding the detector plane, and - discovery (and closure) of various leakage paths through the passive shielding system.

  14. TARGETLESS CAMERA CALIBRATION

    Directory of Open Access Journals (Sweden)

    L. Barazzetti

    2012-09-01

    Full Text Available In photogrammetry a camera is considered calibrated if its interior orientation parameters are known. These encompass the principal distance, the principal point position and some Additional Parameters used to model possible systematic errors. The current state of the art for automated camera calibration relies on the use of coded targets to accurately determine the image correspondences. This paper presents a new methodology for the efficient and rigorous photogrammetric calibration of digital cameras which does not require any longer the use of targets. A set of images depicting a scene with a good texture are sufficient for the extraction of natural corresponding image points. These are automatically matched with feature-based approaches and robust estimation techniques. The successive photogrammetric bundle adjustment retrieves the unknown camera parameters and their theoretical accuracies. Examples, considerations and comparisons with real data and different case studies are illustrated to show the potentialities of the proposed methodology.

  15. Herschel celestial calibration sources: Four large main-belt asteroids as prime flux calibrators for the far-IR/sub-mm range

    CERN Document Server

    Mueller, T G; Nielbock, M; Lim, T; Teyssier, D; Olberg, M; Klaas, U; Linz, H; Altieri, B; Pearson, C; Bendo, G; Vilenius, E

    2013-01-01

    Celestial standards play a major role in observational astrophysics. They are needed to characterise the performance of instruments and are paramount for photometric calibration. During the Herschel Calibration Asteroid Preparatory Programme approximately 50 asteroids have been established as far-IR/sub-mm/mm calibrators for Herschel. The selected asteroids fill the flux gap between the sub-mm/mm calibrators Mars, Uranus and Neptune, and the mid-IR bright calibration stars. All three Herschel instruments observed asteroids for various calibration purposes, including pointing tests, absolute flux calibration, relative spectral response function, observing mode validation, and cross-calibration aspects. Here we present newly established models for the four large and well characterized main-belt asteroids (1) Ceres, (2) Pallas, (4) Vesta, and (21) Lutetia which can be considered as new prime flux calibrators. The relevant object-specific properties (size, shape, spin-properties, albedo, thermal properties) are w...

  16. Absolute spectral radiance responsivity calibration of sun photometers

    Energy Technology Data Exchange (ETDEWEB)

    Xu Qiuyun; Zheng Xiaobing; Zhang Wei; Wang Xianhua; Li Jianjun; Li Xin [Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences, Hefei 230031 (China); Li Zhengqiang [Laboratoire d' Optique Atmospherique, Universite Lille 1, Villeneuve d' Ascq 59655 (France) and State Key Laboratory of Remote Sensing Science, Institute of Remote Sensing Applications, Chinese Academy of Sciences, Beijing 100101 (China)

    2010-03-15

    Sun photometers are designed to measure direct solar irradiance and diffused sky radiance for the purpose of atmospheric parameters characterization. A sun photometer is usually calibrated by using a lamp-illuminated integrating sphere source for its band-averaged radiance responsivity, which normally has an uncertainty of 3%-5% at present. Considering the calibration coefficients may also change with time, a regular high precision calibration is important to maintain data quality. In this paper, a tunable-laser-based facility for spectral radiance responsivity calibration has been developed at the Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences. A reference standard radiance radiometer, calibrated against cryogenic radiometer, is used to determine the radiance from a laser-illuminated integrating sphere source. Spectral radiance responsivity of CIMEL CE318-2 sun photometer is calibrated using this new calibration system with a combined standard uncertainty of about 0.8%. As a validation, the derived band-averaged radiance responsivity are compared to that from a Goddard Space Flight Center lamp-based sphere calibration and good agreements (difference <1.4%) are found from 675 to 1020 nm bands.

  17. Absolute spectral radiance responsivity calibration of sun photometers.

    Science.gov (United States)

    Xu, Qiuyun; Zheng, Xiaobing; Li, Zhengqiang; Zhang, Wei; Wang, Xianhua; Li, Jianjun; Li, Xin

    2010-03-01

    Sun photometers are designed to measure direct solar irradiance and diffused sky radiance for the purpose of atmospheric parameters characterization. A sun photometer is usually calibrated by using a lamp-illuminated integrating sphere source for its band-averaged radiance responsivity, which normally has an uncertainty of 3%-5% at present. Considering the calibration coefficients may also change with time, a regular high precision calibration is important to maintain data quality. In this paper, a tunable-laser-based facility for spectral radiance responsivity calibration has been developed at the Key Laboratory of Optical Calibration and Characterization, Chinese Academy of Sciences. A reference standard radiance radiometer, calibrated against cryogenic radiometer, is used to determine the radiance from a laser-illuminated integrating sphere source. Spectral radiance responsivity of CIMEL CE318-2 sun photometer is calibrated using this new calibration system with a combined standard uncertainty of about 0.8%. As a validation, the derived band-averaged radiance responsivity are compared to that from a Goddard Space Flight Center lamp-based sphere calibration and good agreements (difference <1.4%) are found from 675 to 1020 nm bands.

  18. Landsat-8 Operational Land Imager Radiometric Calibration and Stability

    Directory of Open Access Journals (Sweden)

    Brian Markham

    2014-12-01

    Full Text Available The Landsat-8 Operational Land Imager (OLI was radiometrically calibrated prior to launch in terms of spectral radiance, using an integrating sphere source traceable to National Institute of Standards and Technology (NIST standards of spectral irradiance. It was calibrated on-orbit in terms of reflectance using diffusers characterized prior to launch using NIST traceable standards. The radiance calibration was performed with an uncertainty of ~3%; the reflectance calibration to an uncertainty of ~2%. On-orbit, multiple calibration techniques indicate that the sensor has been stable to better than 0.3% to date, with the exception of the shortest wavelength band, which has degraded about 1.0%. A transfer to orbit experiment conducted using the OLI’s heliostat-illuminated diffuser suggests that some bands increased in sensitivity on transition to orbit by as much as 5%, with an uncertainty of ~2.5%. On-orbit comparisons to other instruments and vicarious calibration techniques show the radiance (without a transfer to orbit adjustment, and reflectance calibrations generally agree with other instruments and ground measurements to within the uncertainties. Calibration coefficients are provided with the data products to convert to either radiance or reflectance units.

  19. Iterative Magnetometer Calibration

    Science.gov (United States)

    Sedlak, Joseph

    2006-01-01

    This paper presents an iterative method for three-axis magnetometer (TAM) calibration that makes use of three existing utilities recently incorporated into the attitude ground support system used at NASA's Goddard Space Flight Center. The method combines attitude-independent and attitude-dependent calibration algorithms with a new spinning spacecraft Kalman filter to solve for biases, scale factors, nonorthogonal corrections to the alignment, and the orthogonal sensor alignment. The method is particularly well-suited to spin-stabilized spacecraft, but may also be useful for three-axis stabilized missions given sufficient data to provide observability.

  20. Retinal vascular calibres are significantly associated with cardiovascular risk factors

    DEFF Research Database (Denmark)

    von Hanno, T.; Bertelsen, G.; Sjølie, Anne K.;

    2014-01-01

    ) and retinal venular calibre (central retinal vein equivalent) were measured computer-assisted on retinal photographs. Data on blood pressure, body mass index (BMI), high-density lipoprotein (HDL) and low-density lipoprotein (LDL) cholesterol, glycosylated haemoglobin (HbA1c) and smoking were collected...... only. Blood pressure had the strongest effect on arteriolar calibre, with a decrease in calibre of 3.6m (women)/4.1m (men) per standard deviation increase in mean arterial blood pressure. Retinal venular calibre was independently associated with age, blood pressure, BMI, HDL and LDL cholesterol....... Association between retinal vessel calibre and the cardiovascular risk factors was assessed by multivariable linear and logistic regression analyses. Results: Retinal arteriolar calibre was independently associated with age, blood pressure, HbA1c and smoking in women and men, and with HDL cholesterol in men...

  1. Calibration of PIXE yields using binary thin films on Si

    Energy Technology Data Exchange (ETDEWEB)

    Meersschaut, J., E-mail: Johan.Meersschaut@imec.be [Imec, Kapeldreef 75, B-3001 Leuven (Belgium); Carbonel, J.; Popovici, M. [Imec, Kapeldreef 75, B-3001 Leuven (Belgium); Zhao, Q.; Vantomme, A. [IKS, KU Leuven, Celestijnenlaan 200D, B-3001 Leuven (Belgium); Vandervorst, W. [Imec, Kapeldreef 75, B-3001 Leuven (Belgium); IKS, KU Leuven, Celestijnenlaan 200D, B-3001 Leuven (Belgium)

    2014-07-15

    We describe the use of binary thin films on Si to calibrate the yields in proton-induced X-ray emission (PIXE) measurements. Besides of the element to be calibrated, the standards also contain a common reference element. The incorporation of a common reference element allows one to eliminate errors in the accumulated beam charge during the calibration of the PIXE set-up. The binary calibration standards allow us to determine the response function with an accuracy close to 1%. As an example, we will perform the calibration for Fe and Co, and we will determine the Co concentration in Fe{sub 1−x}Co{sub x} thin films.

  2. Characterisation of a protection level Am-241 calibration source

    Science.gov (United States)

    Bass, G. A.; Rossiter, M. J.; Williams, T. T.

    1992-11-01

    The various measurements involved in the commissioning process of an Am-241 radioactive source and transport mechanisms to be used for protection level calibration work are detailed. The source and its handling mechanisms are described and measurements to characterize the resultant gamma ray beam are described. For the beam measurements, the inverse square law is investigated and beam uniformity is assessed. A trial calibration of ionization chambers is described. The Am-241 irradiation facility is concluded to be suitable for calibrating secondary standards as part of the calibration service offered for protection level instruments. The umbra part of beam is acceptably uniform for a range of chambers and the measurements obtained were predictable and consistent. This quality will be added to the range of qualities offered as part of the protection level secondary standard calibration service.

  3. The Berkeley extreme ultraviolet calibration facility

    Science.gov (United States)

    Welsh, Barry Y.; Jelinsky, Patrick; Malina, Roger F.

    1988-01-01

    The vacuum calibration facilities of the Space Sciences Laboratory, University of California at Berkeley are designed for the calibration and testing of EUV and FUV spaceborne instrumentation (spectral range 44-2500 A). The facility includes one large cylindrical vacuum chamber (3 x 5 m) containing two EUV collimators, and it is equipped with a 4-axis manipulator of angular-control resolution 1 arcsec for payloads weighing up to 500 kg. In addition, two smaller cylindrical chambers, each 0.9 x 1.2 m, are available for vacuum and thermal testing of UV detectors, filters, and space electronics hardware. All three chambers open into class-10,000 clean rooms, and all calibrations are referred to NBS secondary standards.

  4. A Calibrating Device for Rogowski Coil Development

    Institute of Scientific and Technical Information of China (English)

    LV Liang; LI Junhao; HUANG Jianjun; JI Shengchang; LI Yanming

    2007-01-01

    A calibrating device for the Rogowski coil is developed,which can be used to calibrate the Rogowski coil having a partial response time within tens of nanoseconds.Its key component is a step current generator,which can generate the output with a rise time of less than 2 ns and a duration of larger than 300 ns.The step current generator is composed by a pulse forming line(PFL)and a pulse transmission line(PTL).A TEM(transverse electromagnetic mode)coaxial measurement unit is used as PTL,and the coil to be calibrated and the referenced standard Rogowski coil can be fixed in the unit.The effect of the dimensions of the TEM unit is discussed theoretically as well as experimentally.

  5. Kinematic calibration of orthoglide-type mechanisms

    CERN Document Server

    Pashkevich, Anatoly; Wenger, Philippe

    2006-01-01

    The paper proposes a novel calibration approach for the Orthoglide-type mechanisms based on observations of the manipulator leg parallelism during mo-tions between the prespecified test postures. It employs a low-cost measuring system composed of standard comparator indicators attached to the universal magnetic stands. They are sequentially used for measuring the deviation of the relevant leg location while the manipulator moves the TCP along the Cartesian axes. Using the measured differences, the developed algorithm estimates the joint offsets that are treated as the most essential parameters to be adjusted. The sensitivity of the meas-urement methods and the calibration accuracy are also studied. Experimental re-sults are presented that demonstrate validity of the proposed calibration technique

  6. Laboratory implantation for well type ionization chambers calibration; Implantacao de um laboratorio para calibracao de camaras de ionizacao tipo poco

    Energy Technology Data Exchange (ETDEWEB)

    Vianello, E.A.; Dias, D.J.; Almeida, C.E. de [Laboratorio de Ciencias Radiologicas- LCR- DBB (UERJ). R. Sao Francisco Xavier, 524- Pav. HLC, sala 136 terreo- CEP 20.550-013. Rio de Janeiro (Brazil)

    1998-12-31

    The Radiological Science Laboratory is implanting a service for calibration of well type chambers by IAEA training program. The kerma rate in the air (mu Gy/h) of the linear Cs-137 reference source CDCS-J4 have been determined using a well type chamber Standard Imaging HDR-1000 model, which have been calibrated at Secondary Standard Laboratory Calibration of IAEA, whereas two HDR-1000 Plus chambers were calibrated too, following the same standards. The results were compared with Wisconsin University calibration certification and has demonstrated that well type ionization chamber calibration can be used in brachytherapy for several kinds of radionuclides. (Author)

  7. Numerical simulation of neutron scattering in calibration experiments of a standard 'long counter'; Simulacao numerica do espalhamento em experimentos de calibracao de um medidor padrao 'long Counter'

    Energy Technology Data Exchange (ETDEWEB)

    Caldeira, Alexandre D.; Dias, Artur F.; Claro, Luiz H.; Vieira, Wilson J. [Centro Tecnico Aeroespacial (CTA-IEAv), Sao Jose dos Campos, SP (Brazil). Inst. de Estudos Avancados

    2000-07-01

    This work is part of a research project that involves experimental and numeric theoretical activities of calibration and use of a Long Counter detector in the facilities of the IEAv/CTA electron linear accelerator. The version of the detector considered in this work was based on the project of De Pangher and Nichols to determine the intensity of the sources of fast neutrons that will be produced and to serve as secondary caliper of other neutrons detector in a wide range of energy. The objective is to obtain general information on the scattered neutrons in the experimental atmosphere through preliminary theoretical calculations, using the systems of programs MCNP and DOORS, based on Monte Carlo methods and on the discrete ordinates method, respectively. The results are used to visualize the dimensions of the experimental room, where the influence of neutron scattering in the experimental measurements is the possible smallest. (author)

  8. Uniform calibration of night vision goggles and test sets

    Science.gov (United States)

    Eppeldauer, George P.

    2007-10-01

    There are orders of magnitude differences between the ~0.1 % (k=2) uncertainty of NIST reference detector calibrations and the uncertainty of night vision (NV) goggle measurements. NIST developed a night vision radiometer calibration facility including NV radiometer transfer standards. The transfer standards, that propagate the radiance responsivity scale to the military primary standards laboratories, are calibrated against a NIST reference radiometer. The reference radiometer has been calibrated on the NIST Spectral Comparator Facility (SCF) for spectral power and irradiance responsivities. Spectral considerations are discussed to lower the uncertainties of the radiance responsivity scale transfer to the test sets and then to the goggles. Since direct determination of the final uncertainties in goggle calibrations and measurements is difficult, models have been made to estimate the most important uncertainty components based on individual spectral measurements of the applied source distributions and radiometer spectral responsivities. It is also shown, that because of source spectral mismatch problems, the goggle measurement uncertainty at applications can be much higher than at calibration. A suggestion is being made to mimic the no-moon (stars only) night sky radiation distribution using several LEDs in the test-sets to decrease the large spectral mismatch errors. A broad-band correction factor has been developed to further decrease calibration uncertainty when the goggles to be used have different spectral responsivities than the standard. Geometrical considerations to optimize the radiance measurement angle and the out-of-target blocking are also discussed to decrease the uncertainty in the radiance responsivity transfer.

  9. Coordinate Standard Measurement Development

    Energy Technology Data Exchange (ETDEWEB)

    Hanshaw, R.A.

    2000-02-18

    A Shelton Precision Interferometer Base, which is used for calibration of coordinate standards, was improved through hardware replacement, software geometry error correction, and reduction of vibration effects. Substantial increases in resolution and reliability, as well as reduction in sampling time, were achieved through hardware replacement; vibration effects were reduced substantially through modification of the machine component dampening and software routines; and the majority of the machine's geometry error was corrected through software geometry error correction. Because of these modifications, the uncertainty of coordinate standards calibrated on this device has been reduced dramatically.

  10. Smart Calibration of Excavators

    DEFF Research Database (Denmark)

    Bro, Marie; Døring, Kasper; Ellekilde, Lars-Peter

    2005-01-01

    Excavators dig holes. But where is the bucket? The purpose of this report is to treat four different problems concerning calibrations of position indicators for excavators in operation at concrete construction sites. All four problems are related to the question of how to determine the precise ge...

  11. Calibrating Communication Competencies

    Science.gov (United States)

    Surges Tatum, Donna

    2016-11-01

    The Many-faceted Rasch measurement model is used in the creation of a diagnostic instrument by which communication competencies can be calibrated, the severity of observers/raters can be determined, the ability of speakers measured, and comparisons made between various groups.

  12. Entropic calibration revisited

    Energy Technology Data Exchange (ETDEWEB)

    Brody, Dorje C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)]. E-mail: d.brody@imperial.ac.uk; Buckley, Ian R.C. [Centre for Quantitative Finance, Imperial College, London SW7 2AZ (United Kingdom); Constantinou, Irene C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom); Meister, Bernhard K. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)

    2005-04-11

    The entropic calibration of the risk-neutral density function is effective in recovering the strike dependence of options, but encounters difficulties in determining the relevant greeks. By use of put-call reversal we apply the entropic method to the time reversed economy, which allows us to obtain the spot price dependence of options and the relevant greeks.

  13. Calibration with Absolute Shrinkage

    DEFF Research Database (Denmark)

    Øjelund, Henrik; Madsen, Henrik; Thyregod, Poul

    2001-01-01

    is suggested to cope with the singular design matrix most often seen in chemometric calibration. Furthermore, the proposed algorithm may be generalized to all convex norms like Sigma/beta (j)/(gamma) where gamma greater than or equal to 1, i.e. a method that continuously varies from ridge regression...

  14. LOFAR facet calibration

    CERN Document Server

    van Weeren, R J; Hardcastle, M J; Shimwell, T W; Rafferty, D A; Sabater, J; Heald, G; Sridhar, S S; Dijkema, T J; Brunetti, G; Brüggen, M; Andrade-Santos, F; Ogrean, G A; Röttgering, H J A; Dawson, W A; Forman, W R; de Gasperin, F; Jones, C; Miley, G K; Rudnick, L; Sarazin, C L; Bonafede, A; Best, P N; Bîrzan, L; Cassano, R; Chyży, K T; Croston, J H; Ensslin, T; Ferrari, C; Hoeft, M; Horellou, C; Jarvis, M J; Kraft, R P; Mevius, M; Intema, H T; Murray, S S; Orrú, E; Pizzo, R; Simionescu, A; Stroe, A; van der Tol, S; White, G J

    2016-01-01

    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at $\\sim$ 5arcsec resolu...

  15. Measurement System & Calibration report

    DEFF Research Database (Denmark)

    Kock, Carsten Weber; Vesth, Allan

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  16. Effect of calibration method on Tekscan sensor accuracy.

    Science.gov (United States)

    Brimacombe, Jill M; Wilson, David R; Hodgson, Antony J; Ho, Karen C T; Anglin, Carolyn

    2009-03-01

    Tekscan pressure sensors are used in biomechanics research to measure joint contact loads. While the overall accuracy of these sensors has been reported previously, the effects of different calibration algorithms on sensor accuracy have not been compared. The objectives of this validation study were to determine the most appropriate calibration method supplied in the Tekscan program software and to compare its accuracy to the accuracy obtained with two user-defined calibration protocols. We evaluated the calibration accuracies for test loads within the low range, high range, and full range of the sensor. Our experimental setup used materials representing those found in standard prosthetic joints, i.e., metal against plastic. The Tekscan power calibration was the most accurate of the algorithms provided with the system software, with an overall rms error of 2.7% of the tested sensor range, whereas the linear calibrations resulted in an overall rms error of up to 24% of the tested range. The user-defined ten-point cubic calibration was almost five times more accurate, on average, than the power calibration over the full range, with an overall rms error of 0.6% of the tested range. The user-defined three-point quadratic calibration was almost twice as accurate as the Tekscan power calibration, but was sensitive to the calibration loads used. We recommend that investigators design their own calibration curves not only to improve accuracy but also to understand the range(s) of highest error and to choose the optimal points within the expected sensing range for calibration. Since output and sensor nonlinearity depend on the experimental protocol (sensor type, interface shape and materials, sensor range in use, loading method, etc.), sensor behavior should be investigated for each different application.

  17. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten

    2007-01-01

    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve...... the statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited...

  18. In-situ Broadband Cryogenic Calibration for Two-port Superconducting Microwave Resonators

    CERN Document Server

    Yeh, Jen-Hao

    2012-01-01

    In this paper we introduce an improved microwave calibration method for use in a cryogenic environment, based on a traditional three-standard calibration, the Thru-Reflection-Line (TRL) calibration. The modified calibration method takes advantage of additional information from multiple measurements of an ensemble of realizations of a superconducting resonator, as a new pseudo-Open standard, to correct errors in the TRL calibration. We also demonstrate an experimental realization of this in-situ broadband cryogenic calibration system utilizing cryogenic switches. All calibration measurements are done in the same thermal cycle as the measurement of the resonator (requiring only an additional 20 minutes), thus avoiding 4 additional thermal cycles for traditional TRL calibration (which would require an additional 12 days). The experimental measurements on a wave chaotic microwave billiard verify that the new method significantly improves the measured scattering matrix of a high-quality-factor superconducting reso...

  19. In situ broadband cryogenic calibration for two-port superconducting microwave resonators.

    Science.gov (United States)

    Yeh, Jen-Hao; Anlage, Steven M

    2013-03-01

    We introduce an improved microwave calibration method for use in a cryogenic environment, based on a traditional three-standard calibration, the Thru-Reflect-Line (TRL) calibration. The modified calibration method takes advantage of additional information from multiple measurements of an ensemble of realizations of a superconducting resonator, as a new pseudo-Open standard, to correct errors in the TRL calibration. We also demonstrate an experimental realization of this in situ broadband cryogenic calibration system utilizing cryogenic switches. All calibration measurements are done in the same thermal cycle as the measurement of the resonator (requiring only an additional 20 min), thus avoiding 4 additional thermal cycles for traditional TRL calibration (which would require an additional 12 days). The experimental measurements on a wave-chaotic microwave billiard verify that the new method significantly improves the measured scattering matrix of a high-quality-factor superconducting resonator.

  20. Method for large-range structured light system calibration.

    Science.gov (United States)

    An, Yatong; Bell, Tyler; Li, Beiwen; Xu, Jing; Zhang, Song

    2016-11-20

    Structured light system calibration often requires the usage of a calibration target with a similar size as the field of view (FOV), which brings challenges to a large-range structured light system calibration since fabricating large calibration targets is difficult and expensive. This paper presents a large-range system calibration method that does not need a large calibration target. The proposed method includes two stages: (1) accurately calibrate intrinsics (i.e., focal lengths and principle points) at a near range where both the camera and projector are out of focus, and (2) calibrate the extrinsic parameters (translation and rotation) from camera to projector with the assistance of a low-accuracy, large-range three-dimensional (3D) sensor (e.g., Microsoft Kinect). We have developed a large-scale 3D shape measurement system with a FOV of 1120  mm×1900  mm×1000  mm. Experiments demonstrate our system can achieve measurement accuracy as high as 0.07 mm with a standard deviation of 0.80 mm by measuring a 304.8 mm diameter sphere. As a comparison, Kinect V2 only achieved mean error of 0.80 mm with a standard deviation of 3.41 mm for the FOV of measurement.

  1. Final report on COOMET.AUV.A-S1: Technical report on supplementary comparison 'Comparison of national standards of the sound pressure unit in air through calibration of working reference microphones'

    Science.gov (United States)

    Pozdeeva, Valentina; Chalyy, Vladimir

    2014-01-01

    The supplementary comparison COOMET.AUV.A-S1 for secondary calibration methods using WS1 and WS2 measurement microphones was carried out from 2009 to 2010. The results were submitted to and approved by CCAUV in April 2014. Four National Metrology Institutes took part in this comparison and are as follows: BelGIM (Belarus), VNIIFTRI (Russia), SMU (Slovakia) and DP NDI 'Sistema' (Ukraine). Three of the above NMIs (VNIIFTRI, SMU and DP NDI 'Sistema') had earlier participated in COOMET key comparisons and one NMI (VNIIFTRI) had also participated in CCAUV key comparisons. The Comparison Reference Values were calculated as the weighted mean values from results obtained by three institutes. The comparison results show agreement for all participants in the frequency range from 20 Hz to 12.5 kHz for WS1 microphones, and in the frequency range from 20 Hz to 16 kHz for WS2 microphones. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).

  2. Viscometry for liquids calibration of viscometers

    CERN Document Server

    Gupta, S V

    2014-01-01

    This book is written for scientists involved in the calibration of viscometers. A detailed description for stepping up procedures to establish the viscosity scale and obtaining sets of master viscometers is given in the book. Uncertainty considerations for standard oils of known viscosity are presented. The modern viscometers based on principles of  tuning fork, ultrasonic, PZT, plate waves, Love waves, micro-cantilever and vibration of optical fiber are discussed to inspire the reader to further research and to generate improved versions. The primary standard for viscosity is pure water. Measurements of its viscosity with accuracy/uncertainty achieved are described. The principles of rotational and oscillation viscometers are explained to enhance the knowledge in calibration work. Devices used for specific materials and viscosity in non SI units are discussed with respect to the need to correlate viscosity values obtained by various devices. The description of commercial viscometers meets the needs of the u...

  3. The Calibration Reference Data System

    Science.gov (United States)

    Greenfield, P.; Miller, T.

    2016-07-01

    We describe a software architecture and implementation for using rules to determine which calibration files are appropriate for calibrating a given observation. This new system, the Calibration Reference Data System (CRDS), replaces what had been previously used for the Hubble Space Telescope (HST) calibration pipelines, the Calibration Database System (CDBS). CRDS will be used for the James Webb Space Telescope (JWST) calibration pipelines, and is currently being used for HST calibration pipelines. CRDS can be easily generalized for use in similar applications that need a rules-based system for selecting the appropriate item for a given dataset; we give some examples of such generalizations that will likely be used for JWST. The core functionality of the Calibration Reference Data System is available under an Open Source license. CRDS is briefly contrasted with a sampling of other similar systems used at other observatories.

  4. The Atmospheric Lifetime Experiment. II - Calibration

    Science.gov (United States)

    Rasmussen, R. A.; Lovelock, J. E.

    1983-10-01

    The calibration standards used in the Atmospheric Lifetime Experiment (ALE) for CFCl3, CF2Cl2, CH3CCl3, and CCl4 are described. This includes the preparation of the primary standards by static dilution and their propagation and stability for the period 1977-1982. Two independent assessments of the absolute concentrations of the primary standards used to initiate the ALE measurements in 1977-1978 are reported. For consistency in the ALE program the values assigned to the primary standards and subsequent working standards used in the field were not altered during the experiment when results of better estimates of the original concentration values were obtained. Rather, the appropriate factors by which the ALE mixing ratios for a given species should be multiplied to obtain the best estimate of the current concentration of a given species, are provided.

  5. Lidar calibration experiments

    DEFF Research Database (Denmark)

    Ejsing Jørgensen, Hans; Mikkelsen, T.; Streicher, J.

    1997-01-01

    A series of atmospheric aerosol diffusion experiments combined with lidar detection was conducted to evaluate and calibrate an existing retrieval algorithm for aerosol backscatter lidar systems. The calibration experiments made use of two (almost) identical mini-lidar systems for aerosol cloud...... detection to test the reproducibility and uncertainty of lidars. Lidar data were obtained from both single-ended and double-ended Lidar configurations. A backstop was introduced in one of the experiments and a new method was developed where information obtained from the backstop can be used in the inversion...... algorithm. Independent in-situ aerosol plume concentrations were obtained from a simultaneous tracer gas experiment with SF6, and comparisons with the two lidars were made. The study shows that the reproducibility of the lidars is within 15%, including measurements from both sides of a plume...

  6. Optical tweezers absolute calibration

    CERN Document Server

    Dutra, R S; Neto, P A Maia; Nussenzveig, H M

    2014-01-01

    Optical tweezers are highly versatile laser traps for neutral microparticles, with fundamental applications in physics and in single molecule cell biology. Force measurements are performed by converting the stiffness response to displacement of trapped transparent microspheres, employed as force transducers. Usually, calibration is indirect, by comparison with fluid drag forces. This can lead to discrepancies by sizable factors. Progress achieved in a program aiming at absolute calibration, conducted over the past fifteen years, is briefly reviewed. Here we overcome its last major obstacle, a theoretical overestimation of the peak stiffness, within the most employed range for applications, and we perform experimental validation. The discrepancy is traced to the effect of primary aberrations of the optical system, which are now included in the theory. All required experimental parameters are readily accessible. Astigmatism, the dominant effect, is measured by analyzing reflected images of the focused laser spo...

  7. Newly developed apparatus for calibration of quadrupole mass spectrometer

    Science.gov (United States)

    Meng, Dong; Yongjun, Cheng; Detian, Li; Wenjun, Sun; Lan, Zhao; Meiru, Guo; Yongjun, Wang; Huzhong, Zhang; Yanwu, Li; Gang, Li

    2017-01-01

    In this paper, a new calibration apparatus has been developed for calibrating quadrupole mass spectrometer (QMS) from 10-8 Pa to 10-2 Pa for He, N2 and Ar. Based on the apparatus, two calibration methods are introduced to generate standard pressure P i ‧ (or P i ″) of lower than 10-2 Pa, which are named as ‘the SRG measurement method’ and ‘the CDG measurement method’. The SRG measurement method can generate a standard pressure range from 10-8 Pa to 10-5 Pa, and the CDG measurement method can generate a standard pressure range from 10-5 Pa to 10-2 Pa. A molecular flow condition is determined and a QMS is calibrated by the two methods using single gas and mixture gas. The gas flow through the small tube and orifice realizes molecular flow at an upstream pressure of less than 70 Pa. The sensitivities calibration results with the two methods agree well, and the difference is less than 3%. The sensitivity calibrated with single gas is different from that calibrated with mixture gas. Obtained results are roughly consistent with those of previous studies.

  8. Astrid-2 SSC ASUMagnetic Calibration

    DEFF Research Database (Denmark)

    Primdahl, Fritz

    1997-01-01

    Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory.......Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory....

  9. Calibration Facilities for NIF

    Energy Technology Data Exchange (ETDEWEB)

    Perry, T.S.

    2000-06-15

    The calibration facilities will be dynamic and will change to meet the needs of experiments. Small sources, such as the Manson Source should be available to everyone at any time. Carrying out experiments at Omega is providing ample opportunity for practice in pre-shot preparation. Hopefully, the needs that are demonstrated in these experiments will assure the development of (or keep in service) facilities at each of the laboratories that will be essential for in-house preparation for experiments at NIF.

  10. Mesoscale hybrid calibration artifact

    Science.gov (United States)

    Tran, Hy D.; Claudet, Andre A.; Oliver, Andrew D.

    2010-09-07

    A mesoscale calibration artifact, also called a hybrid artifact, suitable for hybrid dimensional measurement and the method for make the artifact. The hybrid artifact has structural characteristics that make it suitable for dimensional measurement in both vision-based systems and touch-probe-based systems. The hybrid artifact employs the intersection of bulk-micromachined planes to fabricate edges that are sharp to the nanometer level and intersecting planes with crystal-lattice-defined angles.

  11. New in-situ, non-intrusive calibration

    Science.gov (United States)

    Zunino, Heather; Adrian, Ronald; Ding, Liuyang; Prestridge, Kathy

    2014-11-01

    Tomographic particle image velocimetry (PIV) experiments require precise and accurate camera calibration. Standard techniques make assumptions about hard-to-measure camera parameters (i.e. optical axis angle, distortions, etc.)-reducing the calibration accuracy. Additionally, vibrations and slight movements after calibration may cause significant errors-particularly for tomographic PIV. These problems are exacerbated when a calibration target cannot be placed within the test section. A new PIV camera calibration method has been developed to permit precise calibration without placing a calibration target inside the test section or scanning the target over a volume. The method is capable of correcting for dynamic calibration changes occurring between PIV laser pulses. A transparent calibration plate with fine marks on both sides is positioned on the test section window. Dual-plane mapping makes it possible to determine a mapping function containing both position and angular direction of central rays from particles. From this information, central rays can be traced into the test section with high accuracy. Image distortion by the lens and refraction at various air-glass-liquid interfaces are accounted for, and no information about the position or angle of the camera(s) is required.

  12. Effective Calibration of Low-Cost Soil Water Content Sensors

    Directory of Open Access Journals (Sweden)

    Heye Reemt Bogena

    2017-01-01

    Full Text Available Soil water content is a key variable for understanding and modelling ecohydrological processes. Low-cost electromagnetic sensors are increasingly being used to characterize the spatio-temporal dynamics of soil water content, despite the reduced accuracy of such sensors as compared to reference electromagnetic soil water content sensing methods such as time domain reflectometry. Here, we present an effective calibration method to improve the measurement accuracy of low-cost soil water content sensors taking the recently developed SMT100 sensor (Truebner GmbH, Neustadt, Germany as an example. We calibrated the sensor output of more than 700 SMT100 sensors to permittivity using a standard procedure based on five reference media with a known apparent dielectric permittivity (1 < Ka < 34.8. Our results showed that a sensor-specific calibration improved the accuracy of the calibration compared to single “universal” calibration. The associated additional effort in calibrating each sensor individually is relaxed by a dedicated calibration setup that enables the calibration of large numbers of sensors in limited time while minimizing errors in the calibration process.

  13. [Study on the absolute spectral irradiation calibration method for far ultraviolet spectrometer in remote sensing].

    Science.gov (United States)

    Yu, Lei; Lin, Guan-Yu; Chen, Bin

    2013-01-01

    The present paper studied spectral irradiation responsivities calibration method which can be applied to the far ultraviolet spectrometer for upper atmosphere remote sensing. It is difficult to realize the calibration for far ultraviolet spectrometer for many reasons. Standard instruments for far ultraviolet waveband calibration are few, the degree of the vacuum experiment system is required to be high, the stabilities of the experiment are hardly maintained, and the limitation of the far ultraviolet waveband makes traditional diffuser and the integrating sphere radiance calibration method difficult to be used. To solve these problems, a new absolute spectral irradiance calibration method was studied, which can be applied to the far ultraviolet calibration. We build a corresponding special vacuum experiment system to verify the calibration method. The light source system consists of a calibrated deuterium lamp, a vacuum ultraviolet monochromater and a collimating system. We used the calibrated detector to obtain the irradiance responsivities of it. The three instruments compose the calibration irradiance source. We used the "calibration irradiance source" to illuminate the spectrometer prototype and obtained the spectral irradiance responsivities. It realized the absolute spectral irradiance calibration for the far ultraviolet spectrometer utilizing the calibrated detector. The absolute uncertainty of the calibration is 7.7%. The method is significant for the ground irradiation calibration of the far ultraviolet spectrometer in upper atmosphere remote sensing.

  14. Precision Calibration of Infrared Synchrotron Radiation Detectors

    CERN Document Server

    Maltsev, A A; Maslova, M V

    2003-01-01

    The technique of calibration of synchrotron radiation precision detectors on a tungsten source based on similarity (close similarity) of character of spectral distributions of synchrotron and thermal radiations is given. The characteristics of various commonly used lamps, used as "standard" ones, are given. The errors of measurements are analyzed. The detectors are intended for absolute measurements of the number of electrons in a ring-shaped bunch.

  15. Calibration of Underwater Sound Transducers

    Directory of Open Access Journals (Sweden)

    H.R.S. Sastry

    1983-07-01

    Full Text Available The techniques of calibration of underwater sound transducers for farfield, near-field and closed environment conditions are reviewed in this paper .The design of acoustic calibration tank is mentioned. The facilities available at Naval Physical & Oceanographic Laboratory, Cochin for calibration of transducers are also listed.

  16. Comparison of blackbodies for calibration of infrared ear thermometers

    DEFF Research Database (Denmark)

    Pusnik, Igor; Clausen, Sønnik; Favreau, Jacques-Olivier

    2010-01-01

    used for the calibration of IRETs among European national laboratories. To verify the accuracy of an IRET a suitable blackbody is needed. An essential requirement is related to accuracy. There are several standards in the world, which describe requirements for IRETs and a blackbody. In the EU...... this is the standard EN 12470-5, which is a harmonised standard and supports the Medical Device Directive (MDD). Other standards are ASTM standard, Designation E 1965 – 98 and the draft of Japan Industrial Standard JIS T 4207, 2005: Japanese Industrial Standard: “infrared ear thermometers.”. The International...

  17. Calibrating echelle spectrographs with Fabry-Perot etalons

    CERN Document Server

    Bauer, Florian F; Reiners, Ansgar

    2015-01-01

    Over the past decades hollow-cathode lamps have been calibration standards for spectroscopic measurements. Advancing to cm/s radial velocity precisions with the next generation of instruments requires more suitable calibration sources with more lines and less dynamic range problems. Fabry-Perot interferometers provide a regular and dense grid of lines and homogeneous amplitudes making them good candidates for next generation calibrators. We investigate the usefulness of Fabry-Perot etalons in wavelength calibration, present an algorithm to incorporate the etalon spectrum in the wavelength solution and examine potential problems. The quasi periodic pattern of Fabry-Perot lines is used along with a hollow-cathode lamp to anchor the numerous spectral features on an absolute scale. We test our method with the HARPS spectrograph and compare our wavelength solution to the one derived from a laser frequency comb. The combined hollow-cathode lamp/etalon calibration overcomes large distortion (50 m/s) in the wavelengt...

  18. Development of the national radionuclide dose calibrator standardisation service.

    Science.gov (United States)

    van der Gaast, H; Buckman, S; Sherlock, S

    1993-03-01

    The Australian Nuclear Science and Technology Organisation, acting as agent for the CSIRO Division of Applied Physics, maintains the Australian standard of measurement for activity. The standard includes all nuclear medicine gamma emitters and a new standard for pure positron emitters. Under Section 10 of the National Measurement Act 1960, if a measurement is made for a legal purpose, or if the legality of a measurement is in dispute, it can only be confirmed if the following two conditions are fulfilled: (a) that the measurement be in terms of the prescribed Australian legal units of measurement. (b) that it can be proven to be traceable to an Australian primary standard of measurement. To satisfy these requirements, radionuclide dose calibrators require a calibration report determined by Ansto. For this reason, Ansto has developed the national radionuclide dose calibrator standardisation service.

  19. CALIBRATED HYDRODYNAMIC MODEL

    Directory of Open Access Journals (Sweden)

    Sezar Gülbaz

    2015-01-01

    Full Text Available The land development and increase in urbanization in a watershed affect water quantityand water quality. On one hand, urbanization provokes the adjustment of geomorphicstructure of the streams, ultimately raises peak flow rate which causes flood; on theother hand, it diminishes water quality which results in an increase in Total SuspendedSolid (TSS. Consequently, sediment accumulation in downstream of urban areas isobserved which is not preferred for longer life of dams. In order to overcome thesediment accumulation problem in dams, the amount of TSS in streams and inwatersheds should be taken under control. Low Impact Development (LID is a BestManagement Practice (BMP which may be used for this purpose. It is a land planningand engineering design method which is applied in managing storm water runoff inorder to reduce flooding as well as simultaneously improve water quality. LID includestechniques to predict suspended solid loads in surface runoff generated over imperviousurban surfaces. In this study, the impact of LID-BMPs on surface runoff and TSS isinvestigated by employing a calibrated hydrodynamic model for Sazlidere Watershedwhich is located in Istanbul, Turkey. For this purpose, a calibrated hydrodynamicmodel was developed by using Environmental Protection Agency Storm WaterManagement Model (EPA SWMM. For model calibration and validation, we set up arain gauge and a flow meter into the field and obtain rainfall and flow rate data. Andthen, we select several LID types such as retention basins, vegetative swales andpermeable pavement and we obtain their influence on peak flow rate and pollutantbuildup and washoff for TSS. Consequently, we observe the possible effects ofLID on surface runoff and TSS in Sazlidere Watershed.

  20. Rectifying calibration error of Goldmann applanation tonometer is easy!

    Directory of Open Access Journals (Sweden)

    Nikhil S Choudhari

    2014-01-01

    Full Text Available Purpose: Goldmann applanation tonometer (GAT is the current Gold standard tonometer. However, its calibration error is common and can go unnoticed in clinics. Its company repair has limitations. The purpose of this report is to describe a self-taught technique of rectifying calibration error of GAT. Materials and Methods: Twenty-nine slit-lamp-mounted Haag-Streit Goldmann tonometers (Model AT 900 C/M; Haag-Streit, Switzerland were included in this cross-sectional interventional pilot study. The technique of rectification of calibration error of the tonometer involved cleaning and lubrication of the instrument followed by alignment of weights when lubrication alone didn′t suffice. We followed the South East Asia Glaucoma Interest Group′s definition of calibration error tolerance (acceptable GAT calibration error within ±2, ±3 and ±4 mm Hg at the 0, 20 and 60-mm Hg testing levels, respectively. Results: Twelve out of 29 (41.3% GATs were out of calibration. The range of positive and negative calibration error at the clinically most important 20-mm Hg testing level was 0.5 to 20 mm Hg and -0.5 to -18 mm Hg, respectively. Cleaning and lubrication alone sufficed to rectify calibration error of 11 (91.6% faulty instruments. Only one (8.3% faulty GAT required alignment of the counter-weight. Conclusions: Rectification of calibration error of GAT is possible in-house. Cleaning and lubrication of GAT can be carried out even by eye care professionals and may suffice to rectify calibration error in the majority of faulty instruments. Such an exercise may drastically reduce the downtime of the Gold standard tonometer.

  1. Monochromator-Based Absolute Calibration of Radiation Thermometers

    Science.gov (United States)

    Keawprasert, T.; Anhalt, K.; Taubert, D. R.; Hartmann, J.

    2011-08-01

    A monochromator integrating-sphere-based spectral comparator facility has been developed to calibrate standard radiation thermometers in terms of the absolute spectral radiance responsivity, traceable to the PTB cryogenic radiometer. The absolute responsivity calibration has been improved using a 75 W xenon lamp with a reflective mirror and imaging optics to a relative standard uncertainty at the peak wavelength of approximately 0.17 % ( k = 1). Via a relative measurement of the out-of-band responsivity, the spectral responsivity of radiation thermometers can be fully characterized. To verify the calibration accuracy, the absolutely calibrated radiation thermometer is used to measure Au and Cu freezing-point temperatures and then to compare the obtained results with the values obtained by absolute methods, resulting in T - T 90 values of +52 mK and -50 mK for the gold and copper fixed points, respectively.

  2. NVLAP activities at Department of Defense calibration laboratories

    Energy Technology Data Exchange (ETDEWEB)

    Schaeffer, D.M. [Defense Nuclear Agency, Alexandria, VA (United States)

    1993-12-31

    There are 367 active radiological instrument calibration laboratories within the U.S. Department of Defense (DoD). Each of the four services in DoD manages, operates, and certifies the technical proficiency and competency of those laboratories under their cognizance. Each service has designated secondary calibration laboratories to trace all calibration source standards to the National Institute of Standards and Technology. Individual service radiological calibration programs and capabilities, present and future, are described, as well as the measurement quality assurance (MQA) processes for their traceability. National Voluntary Laboratory Accreditation Program (NVLAP) programs for dosimetry systems are briefly summarized. Planned NVLAP accreditation of secondary laboratories is discussed in the context of current technical challenges and future efforts.

  3. High-dose secondary calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Humphreys, J.C. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program.

  4. ALTEA: The instrument calibration

    Energy Technology Data Exchange (ETDEWEB)

    Zaconte, V. [INFN and University of Rome Tor Vergata, Department of Physics, Via della Ricerca Scientifica 1, 00133 Rome (Italy)], E-mail: livio.narici@roma2.infn.it; Belli, F.; Bidoli, V.; Casolino, M.; Di Fino, L.; Narici, L.; Picozza, P.; Rinaldi, A. [INFN and University of Rome Tor Vergata, Department of Physics, Via della Ricerca Scientifica 1, 00133 Rome (Italy); Sannita, W.G. [DISM, University of Genova, Genova (Italy); Department of Psychiatry, SUNY, Stoony Brook, NY (United States); Finetti, N.; Nurzia, G.; Rantucci, E.; Scrimaglio, R.; Segreto, E. [Department of Physics, University and INFN, L' Aquila (Italy); Schardt, D. [GSI/Biophysik, Darmstadt (Germany)

    2008-05-15

    The ALTEA program is an international and multi-disciplinary project aimed at studying particle radiation in space environment and its effects on astronauts' brain functions, as the anomalous perception of light flashes first reported during Apollo missions. The ALTEA space facility includes a 6-silicon telescopes particle detector, and is onboard the International Space Station (ISS) since July 2006. In this paper, the detector calibration at the heavy-ion synchrotron SIS18 at GSI Darmstadt will be presented and compared to the Geant 3 Monte Carlo simulation. Finally, the results of a neural network analysis that was used for ion discrimination on fragmentation data will also be presented.

  5. Dynamic Torque Calibration Unit

    Science.gov (United States)

    Agronin, Michael L.; Marchetto, Carl A.

    1989-01-01

    Proposed dynamic torque calibration unit (DTCU) measures torque in rotary actuator components such as motors, bearings, gear trains, and flex couplings. Unique because designed specifically for testing components under low rates. Measures torque in device under test during controlled steady rotation or oscillation. Rotor oriented vertically, supported by upper angular-contact bearing and lower radial-contact bearing that floats axially to prevent thermal expansion from loading bearings. High-load capacity air bearing available to replace ball bearings when higher load capacity or reduction in rate noise required.

  6. The Calibration Home Base for Imaging Spectrometers

    Directory of Open Access Journals (Sweden)

    Johannes Felix Simon Brachmann

    2016-08-01

    Full Text Available The Calibration Home Base (CHB is an optical laboratory designed for the calibration of imaging spectrometers for the VNIR/SWIR wavelength range. Radiometric, spectral and geometric calibration as well as the characterization of sensor signal dependency on polarization are realized in a precise and highly automated fashion. This allows to carry out a wide range of time consuming measurements in an ecient way. The implementation of ISO 9001 standards in all procedures ensures a traceable quality of results. Spectral measurements in the wavelength range 380–1000 nm are performed to a wavelength uncertainty of +- 0.1 nm, while an uncertainty of +-0.2 nm is reached in the wavelength range 1000 – 2500 nm. Geometric measurements are performed at increments of 1.7 µrad across track and 7.6 µrad along track. Radiometric measurements reach an absolute uncertainty of +-3% (k=1. Sensor artifacts, such as caused by stray light will be characterizable and correctable in the near future. For now, the CHB is suitable for the characterization of pushbroom sensors, spectrometers and cameras. However, it is planned to extend the CHBs capabilities in the near future such that snapshot hyperspectral imagers can be characterized as well. The calibration services of the CHB are open to third party customers from research institutes as well as industry.

  7. NIST Calibration of a Neutron Spectrometer ROSPEC.

    Science.gov (United States)

    Heimbach, Craig

    2006-01-01

    A neutron spectrometer was acquired for use in the measurement of National Institute of Standards and Technology neutron fields. The spectrometer included options for the measurement of low and high energy neutrons, for a total measurement range from 0.01 eV up to 17 MeV. The spectrometer was evaluated in calibration fields and was used to determine the neutron spectrum of an Americium-Beryllium neutron source. The calibration fields used included bare and moderated (252)Cf, monoenergetic neutron fields of 2.5 MeV and 14 MeV, and a thermal-neutron beam. Using the calibration values determined in this exercise, the spectrometer gives a good approximation of the neutron spectrum, and excellent values for neutron fluence, for all NIST calibration fields. The spectrometer also measured an Americium-Beryllium neutron field in a NIST exposure facility and determined the field quite well. The spectrometer measured scattering effects in neutron spectra which previously could be determined only by calculation or integral measurements.

  8. Design, Construction and Calibration of a Solar Radiation Measuring Meter

    OpenAIRE

    Asiegbu, A. Daniel; Echeweozo, E. Odinakachi

    2014-01-01

    A digital solar radiation measuring instrument has been designed, constructed and calibrated. It incorporates a small rectangular silicon photocell as the sensor. On exposure to solar radiation, electromotive force which is proportional to radiation intensity is developed within the circuit. The device correlates voltage developed with available solar intensity. A standard solarimeter was therefore used to calibrate the device to translate the unit of its reading from Volt to Watt per square ...

  9. Calibration for Radiation Protection Equipment for the Measuring Airborne Tritium

    Institute of Scientific and Technical Information of China (English)

    CHEN; Xi-lin; SHEN; En-wei; WEI; Ke-xin; WANG; Kong-zhao; LI; Hou-wen; GE; Jian-an; LV; Xiao-xia

    2012-01-01

    <正>Monitoring airborne tritium is an important routine work in heavy water reactor nuclear power stations and the units related with tritium. Nowadays direct measuring instruments like hand carrying tritium monitors are more often used in routine workshop environment check. Need for calibrating such monitors was suggested. A trial work about the calibration for radiation protection equipment for measuring airborne tritium was carried out with a domestic standard EJ/T 1077-1998 equivalent that of IEC 710.

  10. Generic calibration procedures for nacelle-based profiling lidars

    DEFF Research Database (Denmark)

    Borraccino, Antoine; Courtney, Michael; Wagner, Rozenn

    In power performance testing, it has been demonstrated that the effects of wind speed and direction variations over the rotor disk can no longer be neglected for large wind turbines [1]. A new generation of commercial nacelle-based lidars is now available, offering wind profiling capabilities...... to calibrate profiling nacelle lidars........ Developing standard procedures for power curves using lidars requires assessing lidars measurement uncertainty that is provided by a calibration. Based on the calibration results from two lidars, the Avent 5-beam Demonstrator and the Zephir Dual Mode (ZDM), we present in this paper a generic methodology...

  11. Calibration factors for the SNOOPY - NP-100 neutron dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Moscu, D.F. [Medical Physics and Applied Radiation Sciences, McMaster University, Hamilton, Ontario, L8S 4L8 (Canada)], E-mail: moscudf@mcmaster.ca; McNeill, F.E. [Medical Physics and Applied Radiation Sciences, McMaster University, Hamilton, Ontario, L8S 4L8 (Canada); Chase, J. [Radiation Protection Division, Ontario Power Generation, Whitby, Ontario, L1N 9E3 (Canada)

    2007-10-15

    Within CANDU nuclear power facilities, only a small fraction of workers are exposed to neutron radiation. For these individuals, roughly 4.5% of the total radiation equivalent dose is the result of exposure to neutrons. When this figure is considered across all workers receiving external exposure of any kind, only 0.25% of the total radiation equivalent dose is the result of exposure to neutrons. At many facilities, the NP-100 neutron dosimeter, manufactured by Canberra Industries Incorporated, is employed in both direct and indirect dosimetry methods. Also known as 'SNOOPY', these detectors undergo calibration, which results in a calibration factor relating the neutron count rate to the ambient dose equivalent rate, using a standard Am-Be neutron source. Using measurements presented in a technical note, readings from the dosimeter for six different neutron fields in six source-detector orientations were used, to determine a calibration factor for each of these sources. The calibration factor depends on the neutron energy spectrum and the radiation weighting factor to link neutron fluence to equivalent dose. Although the neutron energy spectra measured in the CANDU workplace are quite different than that of the Am-Be calibration source, the calibration factor remains constant - within acceptable limits - regardless of the neutron source used in the calibration; for the specified calibration orientation and current radiation weighting factors. However, changing the value of the radiation weighting factors would result in changes to the calibration factor. In the event of changes to the radiation weighting factors, it will be necessary to assess whether a change to the calibration process or resulting calibration factor is warranted.

  12. A Simple Accelerometer Calibrator

    Science.gov (United States)

    Salam, R. A.; Islamy, M. R. F.; Munir, M. M.; Latief, H.; Irsyam, M.; Khairurrijal

    2016-08-01

    High possibility of earthquake could lead to the high number of victims caused by it. It also can cause other hazards such as tsunami, landslide, etc. In that case it requires a system that can examine the earthquake occurrence. Some possible system to detect earthquake is by creating a vibration sensor system using accelerometer. However, the output of the system is usually put in the form of acceleration data. Therefore, a calibrator system for accelerometer to sense the vibration is needed. In this study, a simple accelerometer calibrator has been developed using 12 V DC motor, optocoupler, Liquid Crystal Display (LCD) and AVR 328 microcontroller as controller system. The system uses the Pulse Wave Modulation (PWM) form microcontroller to control the motor rotational speed as response to vibration frequency. The frequency of vibration was read by optocoupler and then those data was used as feedback to the system. The results show that the systems could control the rotational speed and the vibration frequencies in accordance with the defined PWM.

  13. 库仑滴定法标定硫代硫酸钠标准溶液%Calibration of Na2S2O3 Standard Solution by Coulometric Titration

    Institute of Scientific and Technical Information of China (English)

    王征帆; 杨艳丽

    2011-01-01

    Concentration of Na2S2O3 standard solution was determined by coulometric titration. During determination, mixed liquid of 1 mol/L KI and 1 mol/L H3PO4 (1 : 1) was used as the electrolyte, I2 from electrolysis process reacted with Na2S2O3, the titration end-point was controlled by dead-stop method. The results show that the determination data with above method is consistent with result of common titration analysis, precision of the method can reach to 0.1%. This method is rapid and accurate,and it can be used to detect Na2S2O3 standard solution concentration instead of the common titration analysis.%利用恒电流库仑滴定法电生碘对NaS2O3标准溶液浓度进行了测定;选择1.0 mol/L KI-1.0 mol/L H3PO4(1:1)溶液混合液作为电解液,电解过程中产生的滴定剂I2与Na2S2O3发生氧化还原反应,滴定终点采用死停终点法控制;该方法测定Na2S2O3结果与常规滴定分析法相吻合,精密度达到0.1%;该方法具有快速,更准确的优点,在化工生产及分析检验工作可代替常规滴定分析用于Na2S2O3标准标准溶液浓度的测定。

  14. Research on the method of calibration for the sulfate hygienic standard by turbidimetry%比浊法测定水中硫酸盐卫生标准检验方法的研究

    Institute of Scientific and Technical Information of China (English)

    董静刚; 汪士蔷

    2011-01-01

    目的:研究GB/T5750.5-2006 1.1硫酸钡比浊法测定饮用水中硫酸盐卫生标准检验方法实验条件.方法:对取样方式、比色波长、比色时间及比色皿宽度等等比色条件进行研究.结果:改进后的方法(λ=360 nm,L=3 cm,10 min后到1 h内比色、用50 ml成套比色管取样等)结果的变异系数小,且标准曲线相关系数也较好,可达到0.9996.结论:改进后的方法扩大了检测的范围、简化了步骤、精密度高准确度好、稳定且线性较好.%Objective:To do the research on experimental conditions of the hygienic standard of sulfate in drinking water by using turbidimetry with GB/T5750. 5 - 2006 1. 1 barium sulfate. Methods: On the sampling method, colorimetric wavelength,time, the width of colorimetric cuvette and other colorimetric conditions. Results: The coefficient of variation is smaller, which is applied by the improved method ( eg: λ =360 nm, L =3 cm, concentrating color in 10 minutes to 1 hour and taking samples with 50 ml colorimetric tube and etc. ) Furthermore, the correlation coefficient of the standard curve is even better, which could be up to 0.9996. Conclusion: The improved method may expand the measuring range and simplified the testing steps so that the result of test can achieve high precision, excellent accuracy as well as good stability and linearity.

  15. Geometric calibration of the circle-plus-arc trajectory.

    Science.gov (United States)

    Hoppe, Stefan; Noo, Frédéric; Dennerlein, Frank; Lauritsch, Günter; Hornegger, Joachim

    2007-12-07

    In this paper, a novel geometric calibration method for C-arm cone-beam scanners is presented which allows the calibration of the circle-plus-arc trajectory. The main idea is the separation of the trajectory into two circular segments (circle segment and arc segment) which are calibrated independently. This separation makes it possible to reuse a calibration phantom which has been successfully applied in clinical environments to calibrate numerous routinely used C-arm systems. For each trajectory segment, the phantom is placed in an optimal position. The two calibration results are then combined by computing the transformation the phantom underwent between the independent calibration runs. This combination can be done in a post-processing step by using standard linear algebra. The method is not limited to circle-plus-arc trajectories and works for any calibration procedure in which the phantom has a preferred orientation with respect to a trajectory segment. Results are presented for both simulated as well as real data acquired with a C-arm system. We also present the first image reconstruction results for the circle-plus-arc trajectory using real C-arm data.

  16. Photometric Calibration of the Supernova Legacy Survey Fields

    CERN Document Server

    Regnault, N; Guy, J; Sullivan, M; Cuillandre, J -C; Astier, P; Balland, C; Basa, S; Carlberg, R G; Fouchez, D; Hardin, D; Hook, I M; Howell, D A; Pain, R; Perrett, K; Pritchet, C J

    2009-01-01

    We present the photometric calibration of the Supernova Legacy Survey (SNLS) fields. The SNLS aims at measuring the distances to SNe Ia at (0.3calibration of the survey dominates the systematic uncertainty of the key measurement of the survey, namely the dark energy equation of state. The photometric calibration of the SNLS requires obtaining a uniform response across the imager, calibrating the science field stars in each survey band (SDSS-like ugriz bands) with respect to standards with known flux in the same bands, and binding the calibration to the UBVRI Landolt standards used to calibrate the nearby SNe from the literature necessary to produce cosmological constraints. The spatial non-uniformities of the imager photometric response are mapped using dithered observations of dense stellar fields. Photometric zero-points against Landolt standards are obtained. The linearity o...

  17. Two-Step Camera Calibration Method Developed for Micro UAV'S

    Science.gov (United States)

    Gašparović, M.; Gajski, D.

    2016-06-01

    The development of unmanned aerial vehicles (UAVs) and continuous price reduction of unmanned systems attracted us to this research. Professional measuring systems are dozens of times more expensive and often heavier than "amateur", non-metric UAVs. For this reason, we tested the DJI Phantom 2 Vision Plus UAV. Phantom's smaller mass and velocity can develop less kinetic energy in relation to the professional measurement platforms, which makes it potentially less dangerous for use in populated areas. In this research, we wanted to investigate the ability of such non-metric UAV and find the procedures under which this kind of UAV may be used for the photogrammetric survey. It is important to emphasize that UAV is equipped with an ultra wide-angle camera with 14MP sensor. Calibration of such cameras is a complex process. In the research, a new two-step process is presented and developed, and the results are compared with standard one-step camera calibration procedure. Two-step process involves initially removed distortion on all images, and then uses these images in the phototriangulation with self-calibration. The paper presents statistical indicators which proved that the proposed two-step process is better and more accurate procedure for calibrating those types of cameras than standard one-step calibration. Also, we suggest two-step calibration process as the standard for ultra-wideangle cameras for unmanned aircraft.

  18. Internal Water Vapor Photoacoustic Calibration

    Science.gov (United States)

    Pilgrim, Jeffrey S.

    2009-01-01

    Water vapor absorption is ubiquitous in the infrared wavelength range where photoacoustic trace gas detectors operate. This technique allows for discontinuous wavelength tuning by temperature-jumping a laser diode from one range to another within a time span suitable for photoacoustic calibration. The use of an internal calibration eliminates the need for external calibrated reference gases. Commercial applications include an improvement of photoacoustic spectrometers in all fields of use.

  19. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities; Metodos de calibracao e de intercomparacao de calibradores de dose utilizados em servicos de medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Costa, Alessandro Martins da

    1999-07-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  20. Probability based calibration of pressure coefficients

    DEFF Research Database (Denmark)

    Hansen, Svend Ole; Pedersen, Marie Louise; Sørensen, John Dalsgaard

    2015-01-01

    not depend on the type of variable action. A probability based calibration of pressure coefficients have been carried out using pressure measurements on the standard CAARC building modelled on scale of 1:383. The extreme pressures measured on the CAARC building model in the wind tunnel have been fitted...... to Gumbel distributions, and these fits are found to represent the data measured with good accuracy. The pressure distributions found have been used in a calibration of partial factors, which should achieve a certain theoretical target reliability index. For a target annual reliability index of 4.......3, the Eurocode partial factor of 1.5 for variable actions agrees well with the inherent uncertainties of wind actions when the pressure coefficients are determined using wind tunnel test results. The increased bias and uncertainty when pressure coefficients mainly are based on structural codes lead to a larger...

  1. Summary of KOMPSAT-5 Calibration and Validation

    Science.gov (United States)

    Yang, D.; Jeong, H.; Lee, S.; Kim, B.

    2013-12-01

    Korean Multi-Purpose Satellite 5 (KOMPSAT-5), equipped with high resolution X-band (9.66 GHz) Synthetic Aperture Radar (SAR), is planning to be launched on August 22, 2013. With the satellite's primary mission objective being providing Geographical Information System (GIS), Ocean monitoring and Land management, and Disaster and ENvironment monitoring (GOLDEN), it is expected that its applications for scientific research on geographical processes will be extensive. In order to meet its mission objective, the KOMPSAT-5 will provide three different kinds of SAR imaging modes; High Resolution Mode (1 m resolution, 5 km swath), Standard Mode (3 m resolution, 30 km swath), and Wide Swath Mode (20 m resolution, 100 km swath). The KOMPSAT-5 will be operated in a 550 km sun-synchronous, dawn- dusk orbit with a 28-day ground repeat cycle providing valuable image information on Earth surface day-or-night and even in bad weather condition. After successful launch of the satellite, it will go through Launch and Early Operation (LEOP) and In-Orbit Testing (IOT) period about for 6 months to carry out various tests on satellite bus and payload systems. The satellite bus system will be tested during the first 3 weeks after the launch focusing on the Attitude and Orbit Control Subsystem (AOCS) and Integrated GPS Occultation Receiver (IGOR) calibration. With the completion of bus system test, the SAR payload system will be calibrated during initial In-Flight check period (11 weeks) by the joint effort of Thales Alenia Space Italy (TAS-I) and Korea Aerospace Research Institute (KARI). The pointing and relative calibration will be carried out during this period by analyzing the doppler frequency and antenna beam pattern of reflected microwave signal from selected regions with uniform backscattering coefficients (e.g. Amazon rainforest). A dedicated SAR calibration, called primary calibration, will be allocated at the end of LEOP for 12 weeks to perform thorough calibration activities

  2. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Kristensen, L.; Jensen, G.; Hansen, A.

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statisticalsignificance of the calibration expressions....... It is concluded that the method has the advantage that many anemometers can be calibrated accurately with a minimum of work and cost. The obvious disadvantage is that the calibration of a set of anemometersmay take more than one month in order to have wind speeds covering a sufficiently large magnitude range...

  3. SURF Model Calibration Strategy

    Energy Technology Data Exchange (ETDEWEB)

    Menikoff, Ralph [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-03-10

    SURF and SURFplus are high explosive reactive burn models for shock initiation and propagation of detonation waves. They are engineering models motivated by the ignition & growth concept of high spots and for SURFplus a second slow reaction for the energy release from carbon clustering. A key feature of the SURF model is that there is a partial decoupling between model parameters and detonation properties. This enables reduced sets of independent parameters to be calibrated sequentially for the initiation and propagation regimes. Here we focus on a methodology for tting the initiation parameters to Pop plot data based on 1-D simulations to compute a numerical Pop plot. In addition, the strategy for tting the remaining parameters for the propagation regime and failure diameter is discussed.

  4. RX130 Robot Calibration

    Science.gov (United States)

    Fugal, Mario

    2012-10-01

    In order to create precision magnets for an experiment at Oak Ridge National Laboratory, a new reverse engineering method has been proposed that uses the magnetic scalar potential to solve for the currents necessary to produce the desired field. To make the magnet it is proposed to use a copper coated G10 form, upon which a drill, mounted on a robotic arm, will carve wires. The accuracy required in the manufacturing of the wires exceeds nominal robot capabilities. However, due to the rigidity as well as the precision servo motor and harmonic gear drivers, there are robots capable of meeting this requirement with proper calibration. Improving the accuracy of an RX130 to be within 35 microns (the accuracy necessary of the wires) is the goal of this project. Using feedback from a displacement sensor, or camera and inverse kinematics it is possible to achieve this accuracy.

  5. Solar Cell Calibration and Measurement Techniques

    Science.gov (United States)

    Bailey, Sheila; Brinker, Dave; Curtis, Henry; Jenkins, Phillip; Scheiman, Dave

    2004-01-01

    The increasing complexity of space solar cells and the increasing international markets for both cells and arrays has resulted in workshops jointly sponsored by NASDA, ESA and NASA. These workshops are designed to obtain international agreement on standardized values for the AMO spectrum and constant, recommend laboratory measurement practices and establish a set of protocols for international comparison of laboratory measurements. A working draft of an ISO standard, WD15387, "Requirements for Measurement and Calibration Procedures for Space Solar Cells" was discussed with a focus on the scope of the document, a definition of primary standard cell, and required error analysis for all measurement techniques. Working groups addressed the issues of Air Mass Zero (AMO) solar constant and spectrum, laboratory measurement techniques, and te international round robin methodology. A summary is presented of the current state of each area and the formulation of the ISO document.

  6. Dose calibrator manufacturer-dependent bias in assays of ¹²³I.

    Science.gov (United States)

    Bergeron, Denis E; Cessna, Jeffrey T; Golas, Daniel B; Young, Rheannan K; Zimmerman, Brian E

    2014-08-01

    Calibration factors for commercial ionization chambers (i.e. dose calibrators) were determined for a solution of (123)I; the activity was based on the 1976 NBS standard. A link between the NIST standard and the International Reference System (SIR) was established. The two major U.S. dose calibrator manufacturers recommend oppositely biased calibration factors, giving a spread of 11.3% in measured activities. With modern quantitative imaging techniques capable of ≤10% accuracy, this bias for a SPECT nuclide is highly significant.

  7. Coincidence corrected efficiency calibration of Compton-suppressed HPGe detectors

    Energy Technology Data Exchange (ETDEWEB)

    Aucott, Timothy [Savannah River Site (SRS), Aiken, SC (United States). Savannah River National Lab. (SRNL); Brand, Alexander [Savannah River Site (SRS), Aiken, SC (United States). Savannah River National Lab. (SRNL); DiPrete, David [Savannah River Site (SRS), Aiken, SC (United States). Savannah River National Lab. (SRNL)

    2015-04-20

    The authors present a reliable method to calibrate the full-energy efficiency and the coincidence correction factors using a commonly-available mixed source gamma standard. This is accomplished by measuring the peak areas from both summing and non-summing decay schemes and simultaneously fitting both the full-energy efficiency, as well as the total efficiency, as functions of energy. By using known decay schemes, these functions can then be used to provide correction factors for other nuclides not included in the calibration standard.

  8. Tectonic calibrations in molecular dating

    Institute of Scientific and Technical Information of China (English)

    Ullasa KODANDARAMAIAH

    2011-01-01

    Molecular dating techniques require the use of calibrations, which are usually fossil or geological vicariance-based.Fossil calibrations have been criticised because they result only in minimum age estimates. Based on a historical biogeographic perspective, Ⅰ suggest that vicariance-based calibrations are more dangerous. Almost all analytical methods in historical biogeography are strongly biased towards inferring vicariance, hence vicariance identified through such methods is unreliable. Other studies, especially of groups found on Gondwanan fragments, have simply assumed vicariance. Although it was previously believed that vicariance was the predominant mode of speciation, mounting evidence now indicates that speciation by dispersal is common, dominating vicariance in several groups. Moreover, the possibility of speciation having occurred before the said geological event cannot be precluded. Thus, geological calibrations can under- or overestimate times, whereas fossil calibrations always result in minimum estimates. Another major drawback of vicariant calibrations is the problem of circular reasoning when the resulting estimates are used to infer ages of biogeographic events. Ⅰ argue that fossil-based dating is a superior alternative to vicariance, primarily because the strongest assumption in the latter, that speciation was caused by the said geological process, is more often than not the most tenuous. When authors prefer to use a combination of fossil and vicariant calibrations, one suggestion is to report results both with and without inclusion of the geological constraints. Relying solely on vicariant calibrations should be strictly avoided.

  9. Cobalt source calibration

    Energy Technology Data Exchange (ETDEWEB)

    Rizvi, H.M.

    1999-12-03

    The data obtained from these tests determine the dose rate of the two cobalt sources in SRTC. Building 774-A houses one of these sources while the other resides in room C-067 of Building 773-A. The data from this experiment shows the following: (1) The dose rate of the No.2 cobalt source in Building 774-A measured 1.073 x 10{sup 5} rad/h (June 17, 1999). The dose rate of the Shepherd Model 109 Gamma cobalt source in Building 773-A measured 9.27 x 10{sup 5} rad/h (June 25, 1999). These rates come from placing the graduated cylinder containing the dosimeter solution in the center of the irradiation chamber. (2) Two calibration tests in the 774-A source placed the graduated cylinder with the dosimeter solution approximately 1.5 inches off center in the axial direction. This movement of the sample reduced the measured dose rate 0.92% from 1.083 x 10{sup 5} rad/h to 1.073 x 10{sup 5} rad/h. and (3) A similar test in the cobalt source in 773-A placed the graduated cylinder approximately 2.0 inches off center in the axial direction. This change in position reduced the measured dose rate by 10.34% from 1.036 x 10{sup 6} to 9.27 x 10{sup 5}. This testing used chemical dosimetry to measure the dose rate of a radioactive source. In this method, one determines the dose by the chemical change that takes place in the dosimeter. For this calibration experiment, the author used a Fricke (ferrous ammonium sulfate) dosimeter. This solution works well for dose rates to 10{sup 7} rad/h. During irradiation of the Fricke dosimeter solution the Fe{sup 2+} ions ionize to Fe{sup 3+}. When this occurs, the solution acquires a slightly darker tint (not visible to the human eye). To determine the magnitude of the change in Fe ions, one places the solution in an UV-VIS Spectrophotometer. The UV-VIS Spectrophotometer measures the absorbency of the solution. Dividing the absorbency by the total time (in minutes) of exposure yields the dose rate.

  10. The Advanced LIGO Photon Calibrators

    CERN Document Server

    Karki, S; Kandhasamy, S; Abbott, B P; Abbott, T D; Anders, E H; Berliner, J; Betzwieser, J; Daveloza, H P; Cahillane, C; Canete, L; Conley, C; Gleason, J R; Goetz, E; Kissel, J S; Izumi, K; Mendell, G; Quetschke, V; Rodruck, M; Sachdev, S; Sadecki, T; Schwinberg, P B; Sottile, A; Wade, M; Weinstein, A J; West, M; Savage, R L

    2016-01-01

    The two interferometers of the Laser Interferometry Gravitaional-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events, and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as Photon Calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO Photon Calibrators that are currently providing fiducial displacements on the order of $10^{-18}$ m/$\\sqrt{\\textrm{Hz}}$ with accuracy and precision of better ...

  11. The Advanced LIGO photon calibrators

    Science.gov (United States)

    Karki, S.; Tuyenbayev, D.; Kandhasamy, S.; Abbott, B. P.; Abbott, T. D.; Anders, E. H.; Berliner, J.; Betzwieser, J.; Cahillane, C.; Canete, L.; Conley, C.; Daveloza, H. P.; De Lillo, N.; Gleason, J. R.; Goetz, E.; Izumi, K.; Kissel, J. S.; Mendell, G.; Quetschke, V.; Rodruck, M.; Sachdev, S.; Sadecki, T.; Schwinberg, P. B.; Sottile, A.; Wade, M.; Weinstein, A. J.; West, M.; Savage, R. L.

    2016-11-01

    The two interferometers of the Laser Interferometry Gravitational-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as photon calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO photon calibrators that are currently providing fiducial displacements on the order of 1 0-18m /√{Hz } with accuracy and precision of better than 1%.

  12. TIME CALIBRATED OSCILLOSCOPE SWEEP CIRCUIT

    Science.gov (United States)

    Smith, V.L.; Carstensen, H.K.

    1959-11-24

    An improved time calibrated sweep circuit is presented, which extends the range of usefulness of conventional oscilloscopes as utilized for time calibrated display applications in accordance with U. S. Patent No. 2,832,002. Principal novelty resides in the provision of a pair of separate signal paths, each of which is phase and amplitude adjustable, to connect a high-frequency calibration oscillator to the output of a sawtooth generator also connected to the respective horizontal deflection plates of an oscilloscope cathode ray tube. The amplitude and phase of the calibration oscillator signals in the two signal paths are adjusted to balance out feedthrough currents capacitively coupled at high frequencies of the calibration oscillator from each horizontal deflection plate to the vertical plates of the cathode ray tube.

  13. Automated calibration of multistatic arrays

    Energy Technology Data Exchange (ETDEWEB)

    Henderer, Bruce

    2017-03-14

    A method is disclosed for calibrating a multistatic array having a plurality of transmitter and receiver pairs spaced from one another along a predetermined path and relative to a plurality of bin locations, and further being spaced at a fixed distance from a stationary calibration implement. A clock reference pulse may be generated, and each of the transmitters and receivers of each said transmitter/receiver pair turned on at a monotonically increasing time delay interval relative to the clock reference pulse. Ones of the transmitters and receivers may be used such that a previously calibrated transmitter or receiver of a given one of the transmitter/receiver pairs is paired with a subsequently un-calibrated one of the transmitters or receivers of an immediately subsequently positioned transmitter/receiver pair, to calibrate the transmitter or receiver of the immediately subsequent transmitter/receiver pair.

  14. Antenna Calibration and Measurement Equipment

    Science.gov (United States)

    Rochblatt, David J.; Cortes, Manuel Vazquez

    2012-01-01

    A document describes the Antenna Calibration & Measurement Equipment (ACME) system that will provide the Deep Space Network (DSN) with instrumentation enabling a trained RF engineer at each complex to perform antenna calibration measurements and to generate antenna calibration data. This data includes continuous-scan auto-bore-based data acquisition with all-sky data gathering in support of 4th order pointing model generation requirements. Other data includes antenna subreflector focus, system noise temperature and tipping curves, antenna efficiency, reports system linearity, and instrument calibration. The ACME system design is based on the on-the-fly (OTF) mapping technique and architecture. ACME has contributed to the improved RF performance of the DSN by approximately a factor of two. It improved the pointing performances of the DSN antennas and productivity of its personnel and calibration engineers.

  15. Novel crystal timing calibration method based on total variation

    Science.gov (United States)

    Yu, Xingjian; Isobe, Takashi; Watanabe, Mitsuo; Liu, Huafeng

    2016-11-01

    A novel crystal timing calibration method based on total variation (TV), abbreviated as ‘TV merge’, has been developed for a high-resolution positron emission tomography (PET) system. The proposed method was developed for a system with a large number of crystals, it can provide timing calibration at the crystal level. In the proposed method, the timing calibration process was formulated as a linear problem. To robustly optimize the timing resolution, a TV constraint was added to the linear equation. Moreover, to solve the computer memory problem associated with the calculation of the timing calibration factors for systems with a large number of crystals, the merge component was used for obtaining the crystal level timing calibration values. Compared with other conventional methods, the data measured from a standard cylindrical phantom filled with a radioisotope solution was sufficient for performing a high-precision crystal-level timing calibration. In this paper, both simulation and experimental studies were performed to demonstrate the effectiveness and robustness of the TV merge method. We compare the timing resolutions of a 22Na point source, which was located in the field of view (FOV) of the brain PET system, with various calibration techniques. After implementing the TV merge method, the timing resolution improved from 3.34 ns at full width at half maximum (FWHM) to 2.31 ns FWHM.

  16. Artifacts for Calibration of Submicron Width Measurements

    Science.gov (United States)

    Grunthaner, Frank; Grunthaner, Paula; Bryson, Charles, III

    2003-01-01

    Artifacts that are fabricated with the help of molecular-beam epitaxy (MBE) are undergoing development for use as dimensional calibration standards with submicron widths. Such standards are needed for calibrating instruments (principally, scanning electron microscopes and scanning probe microscopes) for measuring the widths of features in advanced integrated circuits. Dimensional calibration standards fabricated by an older process that involves lithography and etching of trenches in (110) surfaces of single-crystal silicon are generally reproducible to within dimensional tolerances of about 15 nm. It is anticipated that when the artifacts of the present type are fully developed, their critical dimensions will be reproducible to within 1 nm. These artifacts are expected to find increasing use in the semiconductor-device and integrated- circuit industries as the width tolerances on semiconductor devices shrink to a few nanometers during the next few years. Unlike in the older process, one does not rely on lithography and etching to define the critical dimensions. Instead, one relies on the inherent smoothness and flatness of MBE layers deposited under controlled conditions and defines the critical dimensions as the thicknesses of such layers. An artifact of the present type is fabricated in two stages (see figure): In the first stage, a multilayer epitaxial wafer is grown on a very flat substrate. In the second stage, the wafer is cleaved to expose the layers, then the exposed layers are differentially etched (taking advantage of large differences between the etch rates of the different epitaxial layer materials). The resulting structure includes narrow and well-defined trenches and a shelf with thicknesses determined by the thicknesses of the epitaxial layers from which they were etched. Eventually, it should be possible to add a third fabrication stage in which durable, electronically inert artifacts could be replicated in diamondlike carbon from a master made by

  17. Secondary standard dosimetry laboratory at INFLPR

    Energy Technology Data Exchange (ETDEWEB)

    Scarlat, F.; Minea, R.; Scarisoreanu, A.; Badita, E.; Sima, E.; Dumitrascu, M.; Stancu, E.; Vancea, C., E-mail: scarlat.f@gmail.com [National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest (Romania)

    2011-07-01

    National Institute for Laser, Plasma and Radiation Physics (INFLPR) has constructed a High Energy Secondary Standard Dosimetry Laboratory SSDL-STARDOOR - for performing dosimetric calibrations according to ISO IEC SR/EN 17025:2005 standards. This is outfitted with UNIDOS Secondary Standard Dosimeter from PTW (Freiburg Physikalisch-Technische Werksttaten) calibrated at the PTB-Braunschweig (German Federal Institute of Physics and Metrology). A radiation beam of the quality of Q used by our laboratory as calibration source are provided by INFLPR 7 MeV electron beam linear accelerator mounted in our facility. (author)

  18. A calibrated Franklin chimes

    Science.gov (United States)

    Gonta, Igor; Williams, Earle

    1994-05-01

    Benjamin Franklin devised a simple yet intriguing device to measure electrification in the atmosphere during conditions of foul weather. He constructed a system of bells, one of which was attached to a conductor that was suspended vertically above his house. The device is illustrated in a well-known painting of Franklin (Cohen, 1985). The elevated conductor acquired a potential due to the electric field in the atmosphere and caused a brass ball to oscillate between two bells. The purpose of this study is to extend Franklin's idea by constructing a set of 'chimes' which will operate both in fair and in foul weather conditions. In addition, a mathematical relationship will be established between the frequency of oscillation of a metallic sphere in a simplified geometry and the potential on one plate due to the electrification of the atmosphere. Thus it will be possible to calibrate the 'Franklin Chimes' and to obtain a nearly instantaneous measurement of the potential of the elevated conductor in both fair and foul weather conditions.

  19. Calibration and temperature profile of a tungsten filament lamp

    Energy Technology Data Exchange (ETDEWEB)

    De Izarra, Charles [Groupe de Recherche sur l' Energetique des Milieux Ionises, UMR6606 Universite d' Orleans, CNRS, Faculte des Sciences, Site de Bourges, rue Gaston Berger, BP 4043, 18028 Bourges Cedex (France); Gitton, Jean-Michel, E-mail: Charles.De_Izarra@univ-orleans.f [College Littre, 10 rue Littre, Bourges (France)

    2010-07-15

    The goal of this work proposed for undergraduate students and teachers is the calibration of a tungsten filament lamp from electric measurements that are both simple and precise, allowing to determine the temperature of tungsten filament as a function of the current intensity. This calibration procedure was first applied to a conventional filament lamp (lamp used in automotive lighting) and then tested on a standard tungsten ribbon lamp. The calibration procedure developed was checked by determining the calibration point of the tungsten ribbon lamp with an accuracy of 2%. In addition, for low current intensity, it was observed that the temperature of the filament was not uniform; an explanation is proposed by considering a simple heat transfer model.

  20. Energy dependence evaluation of the patient dose calibrator

    Science.gov (United States)

    Costa, N. A.; Potiens, M. P. A.

    2014-02-01

    The aim of this paper was to evaluate the energy dependence of the kerma-area product meter Patient Dose Calibrator (PDC). The research was done using the calculation of the coefficient calibration of the air kerma values of different energies that are related to the radiation qualities provided on the International Standard of the International Electrotechnical Commission IEC 61267 for conventional X-rays, mammography and computerized tomography (CT), established at the Calibration Laboratory of IPEN. The calibration was made using reference ionization chambers with traceability to the Primary Laboratory Physikalisch-Technische Bundesanstalt (PTB), Germany. The energy dependence of the PDC was determined and the results showed that the PDC has small energy dependence, less than 6%, for the range of recommended energy while to the mammography range the values were 16% and for CT qualities the dependence was 1%.

  1. Ozone measurement systems: associated instrumentation and calibration

    Directory of Open Access Journals (Sweden)

    J. Bellido

    2006-01-01

    Full Text Available The harmful effects produced by ozone have lead to a vast regulation to define and establish the quality goals of ambient air, based on common methods and criteria. The surveillance nets of atmospheric pollution are worldwide extended systems and the applied technology for the ozone measurement is nowadays quite standardized. The aim of this paper is to give a general view of the most common systems used in the ozone measurement in ambient air from a practical point of view. The used instrumentation and the usual calibration methods will be described.

  2. Drift-insensitive distributed calibration of probe microscope scanner in nanometer range: Virtual mode

    Science.gov (United States)

    Lapshin, Rostislav V.

    2016-08-01

    A method of distributed calibration of a probe microscope scanner is suggested. The main idea consists in a search for a net of local calibration coefficients (LCCs) in the process of automatic measurement of a standard surface, whereby each point of the movement space of the scanner can be characterized by a unique set of scale factors. Feature-oriented scanning (FOS) methodology is used as a basis for implementation of the distributed calibration permitting to exclude in situ the negative influence of thermal drift, creep and hysteresis on the obtained results. Possessing the calibration database enables correcting in one procedure all the spatial systematic distortions caused by nonlinearity, nonorthogonality and spurious crosstalk couplings of the microscope scanner piezomanipulators. To provide high precision of spatial measurements in nanometer range, the calibration is carried out using natural standards - constants of crystal lattice. One of the useful modes of the developed calibration method is a virtual mode. In the virtual mode, instead of measurement of a real surface of the standard, the calibration program makes a surface image "measurement" of the standard, which was obtained earlier using conventional raster scanning. The application of the virtual mode permits simulation of the calibration process and detail analysis of raster distortions occurring in both conventional and counter surface scanning. Moreover, the mode allows to estimate the thermal drift and the creep velocities acting while surface scanning. Virtual calibration makes possible automatic characterization of a surface by the method of scanning probe microscopy (SPM).

  3. Calibrating the PAU Survey's 46 Filters

    Science.gov (United States)

    Bauer, A.; Castander, F.; Gaztañaga, E.; Serrano, S.; Sevilla, N.; Tonello, N.; PAU Team

    2016-05-01

    The Physics of the Accelerating Universe (PAU) Survey, being carried out by several Spanish institutions, will image an area of 100-200 square degrees in 6 broad and 40 narrow band optical filters. The team is building a camera (PAUCam) with 18 CCDs, which will be installed in the 4 meter William Herschel Telescope at La Palma in 2013. The narrow band filters will each cover 100Å, with the set spanning 4500-8500Å. The broad band set will consist of standard ugriZy filters. The narrow band filters will provide low-resolution (R˜50) photometric "spectra" for all objects observed in the survey, which will reach a depth of ˜24 mag in the broad bands and ˜22.5 mag (AB) in the narrow bands. Such precision will allow for galaxy photometric redshift errors of 0.0035(1+z), which will facilitate the measurement of cosmological parameters with precision comparable to much larger spectroscopic and photometric surveys. Accurate photometric calibration of the PAU data is vital to the survey's science goals, and is not straightforward due to the large and unusual filter set. We outline the data management pipelines being developed for the survey, both for nightly data reduction and coaddition of multiple epochs, with emphasis on the photometric calibration strategies. We also describe the tools we are developing to test the quality of the reduction and calibration.

  4. A reference material for dynamic displacement calibration

    Science.gov (United States)

    Davighi, A.; Hack, E.; Patterson, E.; Whelan, M.

    2010-06-01

    Calibration of displacement and strain measurement systems is an essential step in providing traceability and confidence in stress and strain distributions obtained from experiment and used to validate simulations employed in engineering design. Reference materials provide a simple, well-defined distribution of the measured quantity that can be traced to an international standard and can be used to assess the uncertainty associated with the measurement system. Previous work has established a reference material and procedure for calibrating optical systems for measuring static, in-plane strain distributions and also demonstrated its use. A new effort is in progress to extend this work to the measurement of three-dimensional displacement distributions induced by cyclic and dynamic loading, including transients and large-scale deformation. The first step in this effort has been to define both the essential and desirable attributes of a reference material for calibrating systems capable of measurements of dynamic displacement and strain. An international consortium of research laboratories, system designers, manufacturers and end-users has identified a list of attributes and members of the experimental mechanics community have been asked to weight the importance of these attributes. The attributes are being utilised to evaluate candidate designs for the reference material which have been generated through a series of brain-storming sessions within the consortium.

  5. Shear calibration biases in weak lensing surveys

    CERN Document Server

    Hirata, C M; Hirata, Christopher M.; Seljak, Uros

    2003-01-01

    We investigate biases induced by the conversion between the observed image shape to shear distortion in current weak lensing analysis methods. Such overall calibration biases cannot be detected by the standard tests such as E/B decomposition or calibration with stars. We find that the non-Gaussianity of point spread function has a significant effect and can lead to up to 15 per cent error on the linear amplitude of fluctuations sigma_8 depending on the method of analysis. This could explain some of the discrepancies seen in recent amplitude determinations from weak lensing. Using an elliptical Laguerre expansion method we develop a re-Gaussianization method which reduces the error to calibration error of order 1 per cent even for poorly resolved galaxies. We also discuss a new type of shear selection bias which results in up to roughly 8 percent underestimation of the signal. It is expected to scale with redshift, inducing errors in the growth factor extraction if not properly corrected for. Understanding and...

  6. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-07-01

    Accurate solar radiation data sets are critical to reducing the expenses associated with mitigating performance risk for solar energy conversion systems, and they help utility planners and grid system operators understand the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of calibration methodologies and the resulting calibration responsivities provided by radiometric calibration service providers such as the National Renewable Energy Laboratory (NREL) and manufacturers of radiometers. Some of these radiometers are calibrated indoors, and some are calibrated outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The reference radiometer calibrations are traceable to the World Radiometric Reference. These different methods of calibration demonstrated 1% to 2% differences in solar irradiance measurement. Analyzing these values will ultimately assist in determining the uncertainties of the radiometer data and will assist in developing consensus on a standard for calibration.

  7. A critical comparison of systematic calibration protocols for activated sludge models: a SWOT analysis.

    Science.gov (United States)

    Sin, Gürkan; Van Hulle, Stijn W H; De Pauw, Dirk J W; van Griensven, Ann; Vanrolleghem, Peter A

    2005-07-01

    Modelling activated sludge systems has gained an increasing momentum after the introduction of activated sludge models (ASMs) in 1987. Application of dynamic models for full-scale systems requires essentially a calibration of the chosen ASM to the case under study. Numerous full-scale model applications have been performed so far which were mostly based on ad hoc approaches and expert knowledge. Further, each modelling study has followed a different calibration approach: e.g. different influent wastewater characterization methods, different kinetic parameter estimation methods, different selection of parameters to be calibrated, different priorities within the calibration steps, etc. In short, there was no standard approach in performing the calibration study, which makes it difficult, if not impossible, to (1) compare different calibrations of ASMs with each other and (2) perform internal quality checks for each calibration study. To address these concerns, systematic calibration protocols have recently been proposed to bring guidance to the modeling of activated sludge systems and in particular to the calibration of full-scale models. In this contribution four existing calibration approaches (BIOMATH, HSG, STOWA and WERF) will be critically discussed using a SWOT (Strengths, Weaknesses, Opportunities, Threats) analysis. It will also be assessed in what way these approaches can be further developed in view of further improving the quality of ASM calibration. In this respect, the potential of automating some steps of the calibration procedure by use of mathematical algorithms is highlighted.

  8. Calibration of High Frequency MEMS Microphones

    Science.gov (United States)

    Shams, Qamar A.; Humphreys, William M.; Bartram, Scott M.; Zuckewar, Allan J.

    2007-01-01

    calibrate these microphones at frequencies up to 80 kHz. The technique relied on the use of a random, ultrasonic broadband centrifugal sound source located in a small anechoic chamber. Phase calibrations of the MEMS microphones were derived from cross spectral phase comparisons between the reference and test substitution microphones and an adjacent and invariant grazing-incidence 1/8-inch standard microphone.

  9. Calibration of the Capintec CRC-712M dose calibrator for (18)F.

    Science.gov (United States)

    Mo, L; Reinhard, M I; Davies, J B; Alexiev, D; Baldock, C

    2006-04-01

    Primary standardisation was performed on a solution of (18)F using the 4pibeta-gamma coincidence counting efficiency-tracing extrapolation method with (60)Co used as a tracer nuclide. The result was used to calibrate the ANSTO secondary standard ionisation chamber which is used to disseminate Australian activity standards for gamma emitters. Using the secondary activity standard for (18)F, the Capintec CRC-712M dose calibrator at the Australian National Medical Cyclotron (NMC) Positron Emission Tomography (PET) Quality Control (QC) Section was calibrated. The dial setting number recommended by the manufacturer for the measurement of the activity of (18)F is 439. In this work, the dial setting numbers for the activity measurement of the solution of (18)F in Wheaton vials were experimentally determined to be 443+/-12, 446+/-12, 459+/-11, 473+/-15 for 0.1, 1, 4.5 and 9ml solution volumes, respectively. The uncertainties given above are expanded uncertainties (k=2) giving an estimated level of confidence of 95%. The activities determined using the manufacturer recommended setting number 439 are 0.8%, 1.4%, 4.0% and 6.5% higher than the standardised activities, respectively. It is recommended that a single dial setting number of 459 determined for 4.5ml is used for 0.1-9ml solution in Wheaton vials in order to simplify the operation procedure. With this setting the expended uncertainty (k=2) in the activity readout from the Capintec dose calibrator would be less than 6.2%.

  10. Mexican national pyronometer network calibration

    Science.gov (United States)

    VAldes, M.; Villarreal, L.; Estevez, H.; Riveros, D.

    2013-12-01

    In order to take advantage of the solar radiation as an alternate energy source it is necessary to evaluate the spatial and temporal availability. The Mexican National Meterological Service (SMN) has a network with 136 meteorological stations, each coupled with a pyronometer for measuring the global solar radiation. Some of these stations had not been calibrated in several years. The Mexican Department of Energy (SENER) in order to count on a reliable evaluation of the solar resource funded this project to calibrate the SMN pyrometer network and validate the data. The calibration of the 136 pyronometers by the intercomparison method recommended by the World Meterological Organization (WMO) requires lengthy observations and specific environmental conditions such as clear skies and a stable atmosphere, circumstances that determine the site and season of the calibration. The Solar Radiation Section of the Instituto de Geofísica of the Universidad Nacional Autónoma de México is a Regional Center of the WMO and is certified to carry out the calibration procedures and emit certificates. We are responsible for the recalibration of the pyronometer network of the SMN. A continuous emission solar simulator with exposed areas with 30cm diameters was acquired to reduce the calibration time and not depend on atmospheric conditions. We present the results of the calibration of 10 thermopile pyronometers and one photovoltaic cell by the intercomparison method with more than 10000 observations each and those obtained with the solar simulator.

  11. Regression calibration with more surrogates than mismeasured variables

    KAUST Repository

    Kipnis, Victor

    2012-06-29

    In a recent paper (Weller EA, Milton DK, Eisen EA, Spiegelman D. Regression calibration for logistic regression with multiple surrogates for one exposure. Journal of Statistical Planning and Inference 2007; 137: 449-461), the authors discussed fitting logistic regression models when a scalar main explanatory variable is measured with error by several surrogates, that is, a situation with more surrogates than variables measured with error. They compared two methods of adjusting for measurement error using a regression calibration approximate model as if it were exact. One is the standard regression calibration approach consisting of substituting an estimated conditional expectation of the true covariate given observed data in the logistic regression. The other is a novel two-stage approach when the logistic regression is fitted to multiple surrogates, and then a linear combination of estimated slopes is formed as the estimate of interest. Applying estimated asymptotic variances for both methods in a single data set with some sensitivity analysis, the authors asserted superiority of their two-stage approach. We investigate this claim in some detail. A troubling aspect of the proposed two-stage method is that, unlike standard regression calibration and a natural form of maximum likelihood, the resulting estimates are not invariant to reparameterization of nuisance parameters in the model. We show, however, that, under the regression calibration approximation, the two-stage method is asymptotically equivalent to a maximum likelihood formulation, and is therefore in theory superior to standard regression calibration. However, our extensive finite-sample simulations in the practically important parameter space where the regression calibration model provides a good approximation failed to uncover such superiority of the two-stage method. We also discuss extensions to different data structures.

  12. Activities of NIST (National Inst. of Standards and Technology)

    Science.gov (United States)

    1992-01-01

    The Radiometric Physics Division of the NIST is responsible for the national standards in radiation thermometry, spectroradiometry, photometry, and spectrophotometry; dissemination of these standards by providing measurement services to customers requiring calibrations of the highest accuracy; and conducting fundamental and applied research to develop the scientific basis for future measurement services. Its relevance to EOS/TIR calibration includes calibrating unknown blackbody for radiance using a well-characterized NIST blackbody source by matching the radiant fluxes with an IR radiometer. The TIR Round Robin is used to verify the calibration of the sources that are used for the absolute radiometric calibration of the individual EOS sensors.

  13. Absolute calibration of optical power for PDT: report of AAPM TG140.

    Science.gov (United States)

    Zhu, Timothy C; Bonnerup, Chris; Colussi, Valdir C; Dowell, Marla L; Finlay, Jarod C; Lilge, Lothar; Slowey, Thomas W; Sibata, Claudio

    2013-08-01

    This report is primarily concerned with methods for optical calibration of laser power for continuous wave (CW) light sources, predominantly used in photodynamic therapy (PDT). Light power calibration is very important for PDT, however, no clear standard has been established for the calibration procedure nor the requirements of power meters suitable for optical power calibration. The purposes of the report are to provide guidance for establishing calibration procedures for thermopile type power meters and establish calibration uncertainties for most commercially available detectors and readout assemblies. The authors have also provided a review of the use of various power meters for CW and pulsed optical sources, and provided recommended temporal frequencies for optical power meter calibrations and guidance for routine quality assurance procedure.

  14. The Landsat Data Continuity Mission Operational Land Imager (OLI) Radiometric Calibration

    Science.gov (United States)

    Markham, Brian L.; Dabney, Philip W.; Murphy-Morris, Jeanine E.; Knight, Edward J.; Kvaran, Geir; Barsi, Julia A.

    2010-01-01

    The Operational Land Imager (OLI) on the Landsat Data Continuity Mission (LDCM) has a comprehensive radiometric characterization and calibration program beginning with the instrument design, and extending through integration and test, on-orbit operations and science data processing. Key instrument design features for radiometric calibration include dual solar diffusers and multi-lamped on-board calibrators. The radiometric calibration transfer procedure from NIST standards has multiple checks on the radiometric scale throughout the process and uses a heliostat as part of the transfer to orbit of the radiometric calibration. On-orbit lunar imaging will be used to track the instruments stability and side slither maneuvers will be used in addition to the solar diffuser to flat field across the thousands of detectors per band. A Calibration Validation Team is continuously involved in the process from design to operations. This team uses an Image Assessment System (IAS), part of the ground system to characterize and calibrate the on-orbit data.

  15. Simple and accurate empirical absolute volume calibration of a multi-sensor fringe projection system

    Science.gov (United States)

    Gdeisat, Munther; Qudeisat, Mohammad; AlSa`d, Mohammed; Burton, David; Lilley, Francis; Ammous, Marwan M. M.

    2016-05-01

    This paper suggests a novel absolute empirical calibration method for a multi-sensor fringe projection system. The optical setup of the projector-camera sensor can be arbitrary. The term absolute calibration here means that the centre of the three dimensional coordinates in the resultant calibrated volume coincides with a preset centre to the three-dimensional real-world coordinate system. The use of a zero-phase fringe marking spot is proposed to increase depth calibration accuracy, where the spot centre is determined with sub-pixel accuracy. Also, a new method is proposed for transversal calibration. Depth and transversal calibration methods have been tested using both single sensor and three-sensor fringe projection systems. The standard deviation of the error produced by this system is 0.25 mm. The calibrated volume produced by this method is 400 mm×400 mm×140 mm.

  16. Jet energy calibration in ATLAS

    CERN Document Server

    Schouten, Doug

    A correct energy calibration for jets is essential to the success of the ATLAS experi- ment. In this thesis I study a method for deriving an in situ jet energy calibration for the ATLAS detector. In particular, I show the applicability of the missing transverse energy projection fraction method. This method is shown to set the correct mean energy for jets. Pileup effects due to the high luminosities at ATLAS are also stud- ied. I study the correlations in lateral distributions of pileup energy, as well as the luminosity dependence of the in situ calibration metho

  17. Calibrating System for Vacuum Gauges

    Institute of Scientific and Technical Information of China (English)

    MengJun; YangXiaotian; HaoBinggan; HouShengjun; HuZhenjun

    2003-01-01

    In order to measure the vacuum degree, a lot of vacuum gauges will be used in CSR vacuum system. We bought several types of vacuum gauges. We know that different typos of vacuum gauges or even one type of vacuum gauges have different measure results in same condition, so they must be calibrated. But it seems impossible for us to send so many gauges to the calibrating station outside because of the high price. So the best choice is to build a second class calibrating station for vacuum gauges by ourselves (Fig.l).

  18. Calibration of ionization chambers in IAEA water phantom; Kalibrering av ionekamre i IAEA-vannfantom

    Energy Technology Data Exchange (ETDEWEB)

    Bjerke, H.; Mikkelborg, O.

    1995-11-01

    The calibration in a water phantom is realised at the Secondary Standard Dosimetry Laboratory in Norway. The method is valid for chambers used in high energy radiation therapy. The method, setup and corrections in the calibration are described. The absorbed dose to air calibration factor, N{sub D}, will have an uncertainty of {approx}0.8% (k=1). 15 refs., 7 figs., 5 tabs.

  19. Improving quantitative precision and throughput by reducing calibrator use in liquid chromatography-tandem mass spectrometry

    Energy Technology Data Exchange (ETDEWEB)

    Rule, Geoffrey S., E-mail: geoffrey.s.rule@aruplab.com [ARUP Institute for Clinical and Experimental Pathology, 500 Chipeta Way, Salt Lake City, UT 84108 (United States); Rockwood, Alan L. [ARUP Institute for Clinical and Experimental Pathology, 500 Chipeta Way, Salt Lake City, UT 84108 (United States); Department of Pathology, University of Utah School of Medicine, 2100 Jones Medical Research Bldg., Salt Lake City, UT 84132 (United States)

    2016-05-05

    To improve efficiency in our mass spectrometry laboratories we have made efforts to reduce the number of calibration standards utilized for quantitation over time. We often analyze three or more batches of 96 samples per day, on a single instrument, for a number of assays. With a conventional calibration scheme at six concentration levels this amounts to more than 5000 calibration points per year. Modern LC-tandem mass spectrometric instrumentation is extremely rugged however, and isotopically labelled internal standards are widely available. This made us consider whether alternative calibration strategies could be utilized to reduce the number of calibration standards analyzed while still retaining high precision and accurate quantitation. Here we demonstrate how, by utilizing a single calibration point in each sample batch, and using the resulting response factor (RF) to update an existing, historical response factor (HRF), we are able to obtain improved precision over a conventional multipoint calibration approach, as judged by quality control samples. The laboratory component of this study was conducted with an existing LC tandem mass spectrometric method for three androgen analytes in our production laboratory. Using examples from both simulated and laboratory data we illustrate several aspects of our single point alternative calibration strategy and compare it with a conventional, multipoint calibration approach. We conclude that both the cost and burden of preparing multiple calibration standards with every batch of samples can be reduced while at the same time maintaining, or even improving, analytical quality. - Highlights: • Use of a weighted single point calibration approach improves quantitative precision. • A weighted response factor approach incorporates historical calibration information. • Several scenarios are discussed with regard to their influence on quantitation.

  20. A kinetic standard for precise calibration of spectrophotometer cell temperature.

    Science.gov (United States)

    Adams, P A; Berman, M C

    1981-05-01

    We describe a simple, highly reproducible kinetic technique for precisely measuring temperature in spectrophotometric systems having reaction cells that are inaccessible to conventional temperature probes. The method is based on the temperature dependence of pseudo-first-order rate constants for the acid-catalyzed hydrolysis of N-o-tolyl-D-glucosylamine. Temperatures of reaction cuvette contents are measured with a precision of +/- 0.05 degrees C (1 SD).

  1. Photometric Calibration of the [$\\alpha$/Fe] Element: I. Calibration with UBV Photometry

    CERN Document Server

    Karaali, S; Bilir, S

    2016-01-01

    We present the calibration of the [$\\alpha$/Fe] element in terms of the ultra-violet excess for 469 dwarf stars with $0.325<(B-V)_0 \\leq 0.775$ mag corresponding the spectral type range F0-K2. The star sample is separated into nine sub-samples with equal range in $(B-V)_0$ colour, $\\Delta(B-V)_0=0.05$ mag, and a third degree polynomial is fitted to each dataset. Our calibrations provide [$\\alpha$/Fe] elements in the range [0.0, 0.4]. We applied the procedure to two sets of field stars and two sets of clusters. The mean and the corresponding standard deviation of the residuals for 43 field stars taken from the Hypatia catalogue are [$\\alpha$/Fe]=-0.090 and $\\sigma =0.102$ dex, while for the 39 ones taken from the same catalogue of stars used in the calibration are [$\\alpha$/Fe]=-0.009 and $\\sigma = 0.079$ dex, respectively. We showed that the differences between the mean of the residuals and standard deviations for two sets of clusters ([$\\alpha$/Fe]=0.073 and $\\sigma = 0.91$ dex; [$\\alpha$/Fe]=-0.012 and $...

  2. Calibration of "Babyline" RP instruments

    CERN Multimedia

    2015-01-01

      If you have old RP instrumentation of the “Babyline” type, as shown in the photo, please contact the Radiation Protection Group (Joffrey Germa, 73171) to have the instrument checked and calibrated. Thank you. Radiation Protection Group

  3. Field calibration of cup anemometers

    Energy Technology Data Exchange (ETDEWEB)

    Kristensen, L.; Jensen, G.; Hansen, A.; Kirkegaard, P.

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statistical significance of the calibration expressions. It is concluded that the method has the advantage that many anemometers can be calibrated accurately with a minimum of work and cost. The obvious disadvantage is that the calibration of a set of anemometers may take more than one month in order to have wind speeds covering a sufficiently large magnitude range in a wind direction sector where we can be sure that the instruments are exposed to identical, simultaneous wind flows. Another main conclusion is that statistical uncertainty must be carefully evaluated since the individual 10 minute wind-speed averages are not statistically independent. (au)

  4. Absolute calibration of an ultraviolet spectrometer using a stabilized laser and a cryogenic cavity radiometer

    Science.gov (United States)

    Jauniskis, L.; Foukal, P.; Kochling, H.

    1992-01-01

    We carry out the calibration of an ultraviolet spectrometer by using a cryogenic electrical-substitution radiometer and intensity-stabilized laser sources. A comparison of the error budgets for the laser-based calibration described here and for a calibration using a type-FEL tungsten spectral-irradiance standard indicates that this technique could provide an improvement of a factor of about three in the uncertainty of the spectrometer calibration, resulting in an absolute accuracy (standard deviation of three) of about 1 percent at 257 nm. The technique described here might significantly improve the accuracy of calibrations on NASA ozone-monitoring and solar ultraviolet-monitoring spectrophotometers when used to complement present procedures that employ lamps and the SURF II synchrotron ultraviolet radiation facility at the National Institute of Standards and Technology.

  5. Calibration of an Automatic System Using a Laser Signature

    Directory of Open Access Journals (Sweden)

    Edward F. Plinski

    2003-04-01

    Full Text Available The specific phenomenon, which appears in tuned CO2 lasers, called a laser signature, is used as a standard for calibration of the servomechanism. The proposed servomechanism can be used for continuous investigations of the laser signatures of different laser media.

  6. Electro-optical equivalent calibration technology for high-energy laser energy meters

    Science.gov (United States)

    Wei, Ji Feng; Chang, Yan; Sun, Li Qun; Zhang, Kai; Hu, Xiao Yang; Zhang, Wei

    2016-04-01

    Electro-optical equivalent calibration with high calibration power and high equivalence is particularly well-suited to the calibration of high-energy laser energy meters. A large amount of energy is reserved during this process, however, which continues to radiate after power-off. This study measured the radiation efficiency of a halogen tungsten lamp during power-on and after power-off in order to calculate the total energy irradiated by a lamp until the high-energy laser energy meter reaches thermal equilibrium. A calibration system was designed based on the measurement results, and the calibration equivalence of the system was analyzed in detail. Results show that measurement precision is significantly affected by the absorption factor of the absorption chamber and by heat loss in the energy meter. Calibration precision is successfully improved by enhancing the equivalent power and reducing power-on time. The electro-optical equivalent calibration system, measurement uncertainty of which was evaluated as 2.4% (k = 2), was used to calibrate a graphite-cone-absorption-cavity absolute energy meter, yielding a calibration coefficient of 1.009 and measurement uncertainty of 3.5% (k = 2). A water-absorption-type high-energy laser energy meter with measurement uncertainty of 4.8% (k = 2) was considered the reference standard, and compared to the energy meter calibrated in this study, yielded a correction factor of 0.995 (standard deviation of 1.4%).

  7. Electro-optical equivalent calibration technology for high-energy laser energy meters.

    Science.gov (United States)

    Wei, Ji Feng; Chang, Yan; Sun, Li Qun; Zhang, Kai; Hu, Xiao Yang; Zhang, Wei

    2016-04-01

    Electro-optical equivalent calibration with high calibration power and high equivalence is particularly well-suited to the calibration of high-energy laser energy meters. A large amount of energy is reserved during this process, however, which continues to radiate after power-off. This study measured the radiation efficiency of a halogen tungsten lamp during power-on and after power-off in order to calculate the total energy irradiated by a lamp until the high-energy laser energy meter reaches thermal equilibrium. A calibration system was designed based on the measurement results, and the calibration equivalence of the system was analyzed in detail. Results show that measurement precision is significantly affected by the absorption factor of the absorption chamber and by heat loss in the energy meter. Calibration precision is successfully improved by enhancing the equivalent power and reducing power-on time. The electro-optical equivalent calibration system, measurement uncertainty of which was evaluated as 2.4% (k = 2), was used to calibrate a graphite-cone-absorption-cavity absolute energy meter, yielding a calibration coefficient of 1.009 and measurement uncertainty of 3.5% (k = 2). A water-absorption-type high-energy laser energy meter with measurement uncertainty of 4.8% (k = 2) was considered the reference standard, and compared to the energy meter calibrated in this study, yielded a correction factor of 0.995 (standard deviation of 1.4%).

  8. Infrasound Sensor Calibration and Response

    Science.gov (United States)

    2012-09-01

    functions with faster rise times. SUMMARY We have documented past work on the determination of the calibration constant of the LANL infrasound sensor...Monitoring Technologies 735 Report Documentation Page Form ApprovedOMB No. 0704-0188 Public reporting burden for the collection of information is estimated...National Laboratory ( LANL ) has operated an infrasound sensor calibration chamber that operates over a frequency range of 0.02 to 4 Hz. This chamber has

  9. Bayesian Calibration of Microsimulation Models.

    Science.gov (United States)

    Rutter, Carolyn M; Miglioretti, Diana L; Savarino, James E

    2009-12-01

    Microsimulation models that describe disease processes synthesize information from multiple sources and can be used to estimate the effects of screening and treatment on cancer incidence and mortality at a population level. These models are characterized by simulation of individual event histories for an idealized population of interest. Microsimulation models are complex and invariably include parameters that are not well informed by existing data. Therefore, a key component of model development is the choice of parameter values. Microsimulation model parameter values are selected to reproduce expected or known results though the process of model calibration. Calibration may be done by perturbing model parameters one at a time or by using a search algorithm. As an alternative, we propose a Bayesian method to calibrate microsimulation models that uses Markov chain Monte Carlo. We show that this approach converges to the target distribution and use a simulation study to demonstrate its finite-sample performance. Although computationally intensive, this approach has several advantages over previously proposed methods, including the use of statistical criteria to select parameter values, simultaneous calibration of multiple parameters to multiple data sources, incorporation of information via prior distributions, description of parameter identifiability, and the ability to obtain interval estimates of model parameters. We develop a microsimulation model for colorectal cancer and use our proposed method to calibrate model parameters. The microsimulation model provides a good fit to the calibration data. We find evidence that some parameters are identified primarily through prior distributions. Our results underscore the need to incorporate multiple sources of variability (i.e., due to calibration data, unknown parameters, and estimated parameters and predicted values) when calibrating and applying microsimulation models.

  10. Pressures Detector Calibration and Measurement

    CERN Document Server

    AUTHOR|(CDS)2156315

    2016-01-01

    This is report of my first and second projects (of 3) in NA61. I did data taking and analysis in order to do calibration of pressure detectors and verified it. I analyzed the data by ROOT software using the C ++ programming language. The first part of my project was determination of calibration factor of pressure sensors. Based on that result, I examined the relation between pressure drop, gas flow rate of in paper filter and its diameter.

  11. Beam Imaging and Luminosity Calibration

    CERN Document Server

    Klute, Markus; Salfeld-Nebgen, Jakob

    2016-01-01

    We discuss a method to reconstruct two-dimensional proton bunch densities using vertex distributions accumulated during LHC beam-beam scans. The $x$-$y$ correlations in the beam shapes are studied and an alternative luminosity calibration technique is introduced. We demonstrate the method on simulated beam-beam scans and estimate the uncertainty on the luminosity calibration associated to the beam-shape reconstruction to be below 1\\%.

  12. Two Methods for Self Calibration of Digital Camera

    Science.gov (United States)

    Sampath, A.; Moe, D.; Christopherson, J.

    2012-07-01

    Photogrammetric mapping using Commercial of the Shelf (COTS) cameras is becoming more popular. Their popularity is augmented by the increasing use of Unmanned Aerial Vehicles (UAV) as a platform for mapping. The mapping precision of these methods can be increased by using a calibrated camera. The USGS/EROS has developed an inexpensive, easy to use method, particularly for calibrating short focal length cameras. The method builds on a self-calibration procedure developed for the USGS EROS Data Center by Pictometry (and augmented by Dr. C.S Fraser), that uses a series of coded targets. These coded targets form different patterns that are imaged from nine different locations with differing camera orientations. A free network solution using collinearity equations is used to determine the calibration parameters. For the smaller focal length COTS cameras, the USGS has developed a procedure that uses a small prototype box that contains these coded targets. The design of the box is discussed, along with best practices for calibration procedure. Results of calibration parameters obtained using the box are compared with the parameters obtained using more established standard procedures.

  13. Calibrating a large slab vessel: A battle of the bulge

    Energy Technology Data Exchange (ETDEWEB)

    Thomas, I.R.

    1993-10-01

    Slab tanks (critically-safe-by-geometry vessels) were proposed for the storage of concentrated, highly-enriched uranium solution in the design of the Fuel Processing Restoration (FPR) Facility at the Idaho Chemical Processing Plant (ICPP). Currently, measurements of bulk mass in ICPP annular vessels have standard deviations on the order of 0.2%, or less. ICPP personnel felt that their inexperience with the aforementioned expansions would prevent them from attaining comparable precision with slab tanks. To help assess the measurement accuracy of slab vessels, a full-scale mockup of those proposed for the FPR Facility was installed for test calibrations. These calibrations were designed to detect vessel expansion under differing conditions. This paper will compare the base-line, water calibrations with those of the higher-density aluminum nitrate, and any observed deflection will be described using vessel calibration techniques. The calibration using water at an elevated temperature was not performed due to the difficulty of maintaining the elevated temperature. This calibration probably will not be conducted because the construction of the FPR Facility has been halted.

  14. Calibration of the Herschel SPIRE Fourier Transform Spectrometer

    CERN Document Server

    Swinyard, B M; Hopwood, R; Valtchanov, I; Lu, N; Fulton, T; Benielli, D; Imhof, P; Marchili, N; Baluteau, J -P; Bendo, G J; Ferlet, M; Griffin, M J; Lim, T L; Makiwa, G; Naylor, D A; Orton, G S; Papageorgiou, A; Pearson, C P; Schulz, B; Sidher, S D; Spencer, L D; van der Wiel, M H D; Wu, R

    2014-01-01

    The Herschel SPIRE instrument consists of an imaging photometric camera and an imaging Fourier Transform Spectrometer (FTS), both operating over a frequency range of 450-1550 GHz. In this paper, we briefly review the FTS design, operation, and data reduction, and describe in detail the approach taken to relative calibration (removal of instrument signatures) and absolute calibration against standard astronomical sources. The calibration scheme assumes a spatially extended source and uses the Herschel telescope as primary calibrator. Conversion from extended to point-source calibration is carried out using observations of the planet Uranus. The model of the telescope emission is shown to be accurate to within 6% and repeatable to better than 0.06% and, by comparison with models of Mars and Neptune, the Uranus model is shown to be accurate to within 3%. Multiple observations of a number of point-like sources show that the repeatability of the calibration is better than 1%, if the effects of the satellite absolu...

  15. Data filtering with support vector machines in geometric camera calibration.

    Science.gov (United States)

    Ergun, B; Kavzoglu, T; Colkesen, I; Sahin, C

    2010-02-01

    The use of non-metric digital cameras in close-range photogrammetric applications and machine vision has become a popular research agenda. Being an essential component of photogrammetric evaluation, camera calibration is a crucial stage for non-metric cameras. Therefore, accurate camera calibration and orientation procedures have become prerequisites for the extraction of precise and reliable 3D metric information from images. The lack of accurate inner orientation parameters can lead to unreliable results in the photogrammetric process. A camera can be well defined with its principal distance, principal point offset and lens distortion parameters. Different camera models have been formulated and used in close-range photogrammetry, but generally sensor orientation and calibration is performed with a perspective geometrical model by means of the bundle adjustment. In this study, support vector machines (SVMs) using radial basis function kernel is employed to model the distortions measured for Olympus Aspherical Zoom lens Olympus E10 camera system that are later used in the geometric calibration process. It is intended to introduce an alternative approach for the on-the-job photogrammetric calibration stage. Experimental results for DSLR camera with three focal length settings (9, 18 and 36 mm) were estimated using bundle adjustment with additional parameters, and analyses were conducted based on object point discrepancies and standard errors. Results show the robustness of the SVMs approach on the correction of image coordinates by modelling total distortions on-the-job calibration process using limited number of images.

  16. 40 CFR 91.424 - Dilute sampling procedure-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... for Standards and Testing (NIST), and will serve as the reference value (NIST “true” value) for the... approved in advance by the Administrator. (b) After the calibration curve has been obtained, verification...., analyzer calibration, leaks, or HC hangup). A verification procedure is found in paragraph (e) of...

  17. Clusters of Monoisotopic Elements for Calibration in (TOF) Mass Spectrometry

    Science.gov (United States)

    Kolářová, Lenka; Prokeš, Lubomír; Kučera, Lukáš; Hampl, Aleš; Peňa-Méndez, Eladia; Vaňhara, Petr; Havel, Josef

    2016-12-01

    Precise calibration in TOF MS requires suitable and reliable standards, which are not always available for high masses. We evaluated inorganic clusters of the monoisotopic elements gold and phosphorus (Au n +/Au n - and P n +/P n -) as an alternative to peptides or proteins for the external and internal calibration of mass spectra in various experimental and instrumental scenarios. Monoisotopic gold or phosphorus clusters can be easily generated in situ from suitable precursors by laser desorption/ionization (LDI) or matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS). Their use offers numerous advantages, including simplicity of preparation, biological inertness, and exact mass determination even at lower mass resolution. We used citrate-stabilized gold nanoparticles to generate gold calibration clusters, and red phosphorus powder to generate phosphorus clusters. Both elements can be added to samples to perform internal calibration up to mass-to-charge (m/z) 10-15,000 without significantly interfering with the analyte. We demonstrated the use of the gold and phosphorous clusters in the MS analysis of complex biological samples, including microbial standards and total extracts of mouse embryonic fibroblasts. We believe that clusters of monoisotopic elements could be used as generally applicable calibrants for complex biological samples.

  18. Study of glass hydrometer calibration by hydrostatic weighting

    Science.gov (United States)

    Chen, Chaoyun; Wang, Jintao; Li, Zhihao; Zhang, Peiman

    2016-01-01

    Glass hydrometers are simple but effective instruments for measuring the density of liquids. Glass hydrometers calibration based on the Archimedes law, using silicon ring as a reference standard solid density, n-tridecane with density stability and low surface tension as the standard working liquid, based on hydrostatic weighing method designs a glass hydrometer calibration system. Glass hydrometer calibration system uses CCD image measurement system to align the scale of hydrometer and liquid surface, with positioning accuracy of 0.01 mm. Surface tension of the working liquid is measured by Whihemy plate. According to twice glass hydrometer weighing in the air and liquid can calculate the correction value of the current scale. In order to verify the validity of the principle of the hydrostatic weighing method of glass hydrometer calibration system, for measuring the density range of (770-790) kg/m3, with a resolution of 0.2 kg/m3 of hydrometer. The results of measurement compare with the Physikalisch-Technische Bundesanstalt(PTB) ,verifying the validity of the calibration system.

  19. Consequences of Secondary Calibrations on Divergence Time Estimates.

    Directory of Open Access Journals (Sweden)

    John J Schenk

    Full Text Available Secondary calibrations (calibrations based on the results of previous molecular dating studies are commonly applied in divergence time analyses in groups that lack fossil data; however, the consequences of applying secondary calibrations in a relaxed-clock approach are not fully understood. I tested whether applying the posterior estimate from a primary study as a prior distribution in a secondary study results in consistent age and uncertainty estimates. I compared age estimates from simulations with 100 randomly replicated secondary trees. On average, the 95% credible intervals of node ages for secondary estimates were significantly younger and narrower than primary estimates. The primary and secondary age estimates were significantly different in 97% of the replicates after Bonferroni corrections. Greater error in magnitude was associated with deeper than shallower nodes, but the opposite was found when standardized by median node age, and a significant positive relationship was determined between the number of tips/age of secondary trees and the total amount of error. When two secondary calibrated nodes were analyzed, estimates remained significantly different, and although the minimum and median estimates were associated with less error, maximum age estimates and credible interval widths had greater error. The shape of the prior also influenced error, in which applying a normal, rather than uniform, prior distribution resulted in greater error. Secondary calibrations, in summary, lead to a false impression of precision and the distribution of age estimates shift away from those that would be inferred by the primary analysis. These results suggest that secondary calibrations should not be applied as the only source of calibration in divergence time analyses that test time-dependent hypotheses until the additional error associated with secondary calibrations is more properly modeled to take into account increased uncertainty in age estimates.

  20. A Method to Test Model Calibration Techniques: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, Ron; Polly, Ben; Neymark, Joel

    2016-09-01

    This paper describes a method for testing model calibration techniques. Calibration is commonly used in conjunction with energy retrofit audit models. An audit is conducted to gather information about the building needed to assemble an input file for a building energy modeling tool. A calibration technique is used to reconcile model predictions with utility data, and then the 'calibrated model' is used to predict energy savings from a variety of retrofit measures and combinations thereof. Current standards and guidelines such as BPI-2400 and ASHRAE-14 set criteria for 'goodness of fit' and assume that if the criteria are met, then the calibration technique is acceptable. While it is logical to use the actual performance data of the building to tune the model, it is not certain that a good fit will result in a model that better predicts post-retrofit energy savings. Therefore, the basic idea here is that the simulation program (intended for use with the calibration technique) is used to generate surrogate utility bill data and retrofit energy savings data against which the calibration technique can be tested. This provides three figures of merit for testing a calibration technique, 1) accuracy of the post-retrofit energy savings prediction, 2) closure on the 'true' input parameter values, and 3) goodness of fit to the utility bill data. The paper will also discuss the pros and cons of using this synthetic surrogate data approach versus trying to use real data sets of actual buildings.

  1. Fusion yield measurements on JET and their calibration

    Energy Technology Data Exchange (ETDEWEB)

    Syme, D.B., E-mail: brian.syme@ccfe.ac.uk [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Popovichev, S. [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Conroy, S. [EURATOM-VR Association, Department of Physics and Astronomy, Uppsala University, Box 516, SE-75120 Uppsala (Sweden); Lengar, I.; Snoj, L. [EURATOM-MHEST Association, Reactor Physics Division, Jožef Stefan Institute, Jamova cesta 39, SI-1000 Ljubljana (Slovenia); Sowden, C. [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Giacomelli, L. [EURATOM-ENEA-CNR Association, CNR-IFP and Univ. di Milano-Bicocca, Milan (Italy); Hermon, G.; Allan, P.; Macheta, P.; Plummer, D.; Stephens, J. [EURATOM-CCFE Fusion Association, Culham Science Centre, Abingdon, OXON OX14 3DB (United Kingdom); Batistoni, P. [EURATOM-ENEA Association, Via E. Fermi,40, 00044 Frascati (Italy); Prokopowicz, R.; Jednorog, S. [EURATOM-IPPLM Association, Institute of Plasma Physics and Laser Microfusion, Hery 23, 01-497 Warsaw (Poland); Abhangi, M.R.; Makwana, R. [Institute for Plasma Research, Bhat, Gandhinagar, 382 428 Gujarat (India)

    2014-11-15

    The power output of fusion experiments and fusion reactor-like devices is measured in terms of the neutron yields which relate directly to the fusion yield. In this paper we describe the devices and methods used to make the new in situ calibration of JET in April 2013 and its early results. The target accuracy of this calibration was 10%, just as in the earlier JET calibration and as required for ITER, where a precise neutron yield measurement is important, e.g., for tritium accountancy. We discuss the constraints and early decisions which defined the main calibration approach, e.g., the choice of source type and the deployment method. We describe the physics, source issues, safety and engineering aspects required to calibrate directly the Fission Chambers and the Activation System which carry the JET neutron calibration. In particular a direct calibration of the Activation system was planned for the first time in JET. We used the existing JET remote-handling system to deploy the {sup 252}Cf source and developed the compatible tooling and systems necessary to ensure safe and efficient deployment in these cases. The scientific programme has sought to better understand the limitations of the calibration, to optimise the measurements and other provisions, to provide corrections for perturbing factors (e.g., presence of the remote-handling boom and other non-standard torus conditions) and to ensure personnel safety and safe working conditions. Much of this work has been based on an extensive programme of Monte-Carlo calculations which, e.g., revealed a potential contribution to the neutron yield via a direct line of sight through the ports which presents individually depending on the details of the port geometry.

  2. Hierarchical Bayesian Data Analysis in Radiometric SAR System Calibration: A Case Study on Transponder Calibration with RADARSAT-2 Data

    Directory of Open Access Journals (Sweden)

    Björn J. Döring

    2013-12-01

    Full Text Available A synthetic aperture radar (SAR system requires external absolute calibration so that radiometric measurements can be exploited in numerous scientific and commercial applications. Besides estimating a calibration factor, metrological standards also demand the derivation of a respective calibration uncertainty. This uncertainty is currently not systematically determined. Here for the first time it is proposed to use hierarchical modeling and Bayesian statistics as a consistent method for handling and analyzing the hierarchical data typically acquired during external calibration campaigns. Through the use of Markov chain Monte Carlo simulations, a joint posterior probability can be conveniently derived from measurement data despite the necessary grouping of data samples. The applicability of the method is demonstrated through a case study: The radar reflectivity of DLR’s new C-band Kalibri transponder is derived through a series of RADARSAT-2 acquisitions and a comparison with reference point targets (corner reflectors. The systematic derivation of calibration uncertainties is seen as an important step toward traceable radiometric calibration of synthetic aperture radars.

  3. ODX - diagnostic standard and development tools; ODX Diagnosestandard und Entwicklungswerkzeuge

    Energy Technology Data Exchange (ETDEWEB)

    Gomez, A.; Kricke, C.; Meyer, J. [ETAS GmbH, Stuttgart (Germany)

    2005-09-01

    ODX, the new standard for the description of diagnostic protocols and data for electronic control units was released a year ago. After a brief review of the standard, ETAS introduces development tools for measuring, calibration, and diagnostic applications, with an emphasis on ECU calibration tasks. (orig.)

  4. Semi-empirical correction of ab initio harmonic properties by scaling factors: a validated uncertainty model for calibration and prediction

    CERN Document Server

    Pernot, Pascal

    2010-01-01

    Bayesian Model Calibration is used to revisit the problem of scaling factor calibration for semi-empirical correction of ab initio harmonic properties (e.g. vibrational frequencies and zero-point energies). A particular attention is devoted to the evaluation of scaling factor uncertainty, and to its effect on the accuracy of scaled properties. We argue that in most cases of interest the standard calibration model is not statistically valid, in the sense that it is not able to fit experimental calibration data within their uncertainty limits. This impairs any attempt to use the results of the standard model for uncertainty analysis and/or uncertainty propagation. We propose to include a stochastic term in the calibration model to account for model inadequacy. This new model is validated in the Bayesian Model Calibration framework. We provide explicit formulae for prediction uncertainty in typical limit cases: large and small calibration sets of data with negligible measurement uncertainty, and datasets with la...

  5. Using LabVIEW to facilitate calibration and verification for respiratory impedance plethysmography.

    Science.gov (United States)

    Ellis, W S; Jones, R T

    1991-12-01

    A system for calibrating the Respitrace impedance plethysmograph was developed with the capacity to quantitatively verify the accuracy of calibration. LabVIEW software was used on a Macintosh II computer to create a user-friendly environment, with the added benefit of reducing development time. The system developed enabled a research assistant to calibrate the Respitrace within 15 min while achieving an accuracy within the normally accepted 10% deviation when the Respitrace output is compared to a water spirometer standard. The system and methods described were successfully used in a study of 10 subjects smoking cigarettes containing marijuana or cocaine under four conditions, calibrating all subjects to 10% accuracy within 15 min.

  6. A least squares procedure for calculating the calibration constants of a portable gamma-ray spectrometer.

    Science.gov (United States)

    Ribeiro, F B; Carlos, D U; Hiodo, F Y; Strobino, E F

    2005-01-01

    In this study, a least squares procedure for calculating the calibration constants of a portable gamma-ray spectrometer using the general inverse matrix method is presented. The procedure weights the model equations fitting to the calibration data, taking into account the variances in the counting rates and in the radioactive standard concentrations. The application of the described procedure is illustrated by calibrating twice the same gamma-ray spectrometer, with two independent data sets collected approximately 18 months apart in the same calibration facility.

  7. The KLOE Online Calibration System

    Institute of Scientific and Technical Information of China (English)

    E.Pasqualucci

    2001-01-01

    Based on all the features of the KLOE online software,the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed.Acalibration manager process controls the system,implementing the interface to the online system,receiving information from the run control and translating its state transitions to a separate state machine.It acts as a " calibration run controller"and performs failure recovery when requested by a set of process checkers.The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms.A client library and C,fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an bool-like interface.Several calibration processes running in parallel in a destributed,multiplatform environment can fill the same histograms,allowing fast external information check.A monitor thread allow remote browsing for visual inspection,Pre-filtered data are read in nonprivileged spy mode from the data acquisition system via the Kloe Integrated Dataflow,privileged spy mode from the data acquisiton system via the Kole Integrated Dataflow.The main characteristics of the system are presented.

  8. Auto-calibration of Systematic Odometry Errors in Mobile Robots

    DEFF Research Database (Denmark)

    Bak, Martin; Larsen, Thomas Dall; Andersen, Nils Axel;

    1999-01-01

    This paper describes the phenomenon of systematic errors in odometry models in mobile robots and looks at various ways of avoiding it by means of auto-calibration. The systematic errors considered are incorrect knowledge of the wheel base and the gains from encoder readings to wheel displacement....... By auto-calibration we mean a standardized procedure which estimates the uncertainties using only on-board equipment such as encoders, an absolute measurement system and filters; no intervention by operator or off-line data processing is necessary. Results are illustrated by a number of simulations...

  9. Improved photon counting efficiency calibration using superconducting single photon detectors

    Science.gov (United States)

    Gan, Haiyong; Xu, Nan; Li, Jianwei; Sun, Ruoduan; Feng, Guojin; Wang, Yanfei; Ma, Chong; Lin, Yandong; Zhang, Labao; Kang, Lin; Chen, Jian; Wu, Peiheng

    2015-10-01

    The quantum efficiency of photon counters can be measured with standard uncertainty below 1% level using correlated photon pairs generated through spontaneous parametric down-conversion process. Normally a laser in UV, blue or green wavelength range with sufficient photon energy is applied to produce energy and momentum conserved photon pairs in two channels with desired wavelengths for calibration. One channel is used as the heralding trigger, and the other is used for the calibration of the detector under test. A superconducting nanowire single photon detector with advantages such as high photon counting speed (optical spectroscopy, super resolution microscopy, deep space observation, and so on.

  10. Calibration of Frequency Data Collection Systems Using Shortwave Radio Signals

    Science.gov (United States)

    Estler, Ron

    2000-09-01

    The atomic-clock-derived audio tones broadcast on the National Institute of Standards and Technology (NIST) shortwave station WWV are used to calibrate computer frequency data collection systems via Fast Fourier Transforms (FFT). Once calibrated, the data collection system can be used to accurately determine the audio signals used in several instructional physical chemistry laboratory experiments. This method can be applied to virtually any hardware-software configuration that allows adjustment of the apparent time scale (digitizing rate) of the recorded audio file.

  11. Calibration and investigation of infrared camera systems applying blackbody radiation

    Science.gov (United States)

    Hartmann, Juergen; Fischer, Joachim

    2001-03-01

    An experimental facility is presented, which allows calibration and detailed investigation of infrared camera systems. Various blackbodies operating in the temperature range from -60 degree(s)C up to 3000 degree(s)C serve as standard radiation sources, enabling calibration of camera systems in a wide temperature and spectral range with highest accuracy. Quantitative results and precise long-term investigations, especially in detecting climatic trends, require accurate traceability to the International Temperature Scale of 1990 (ITS-90). For the used blackbodies the traceability to ITS- 90 is either realized by standard platinum resistance thermometers (in the temperature range below 962 degree(s)C) or by absolute and relative radiometry (in the temperature range above 962 degree(s)C). This traceability is fundamental for implementation of quality assurance systems and realization of different standardizations, for example according ISO 9000. For investigation of the angular and the temperature resolution our set-up enables minimum resolvable (MRTD) and minimum detectable temperature difference (MDTD) measurements in the various temperature ranges. A collimator system may be used to image the MRTD and MDTD targets to infinity. As internal calibration of infrared camera systems critically depends on the temperature of the surrounding, the calibration and investigation of the cameras is performed in a climate box, which allows a detailed controlling of the environmental parameters like humidity and temperature. Experimental results obtained for different camera systems are presented and discussed.

  12. Court Reconstruction for Camera Calibration in Broadcast Basketball Videos.

    Science.gov (United States)

    Wen, Pei-Chih; Cheng, Wei-Chih; Wang, Yu-Shuen; Chu, Hung-Kuo; Tang, Nick C; Liao, Hong-Yuan Mark

    2016-05-01

    We introduce a technique of calibrating camera motions in basketball videos. Our method particularly transforms player positions to standard basketball court coordinates and enables applications such as tactical analysis and semantic basketball video retrieval. To achieve a robust calibration, we reconstruct the panoramic basketball court from a video, followed by warping the panoramic court to a standard one. As opposed to previous approaches, which individually detect the court lines and corners of each video frame, our technique considers all video frames simultaneously to achieve calibration; hence, it is robust to illumination changes and player occlusions. To demonstrate the feasibility of our technique, we present a stroke-based system that allows users to retrieve basketball videos. Our system tracks player trajectories from broadcast basketball videos. It then rectifies the trajectories to a standard basketball court by using our camera calibration method. Consequently, users can apply stroke queries to indicate how the players move in gameplay during retrieval. The main advantage of this interface is an explicit query of basketball videos so that unwanted outcomes can be prevented. We show the results in Figs. 1, 7, 9, 10 and our accompanying video to exhibit the feasibility of our technique.

  13. Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Faber, M.H.; Sørensen, John Dalsgaard

    2003-01-01

    The present paper addresses fundamental concepts of reliability based code calibration. First basic principles of structural reliability theory are introduced and it is shown how the results of FORM based reliability analysis may be related to partial safety factors and characteristic values....... Thereafter the code calibration problem is presented in its principal decision theoretical form and it is discussed how acceptable levels of failure probability (or target reliabilities) may be established. Furthermore suggested values for acceptable annual failure probabilities are given for ultimate...... and serviceability limit states. Finally the paper describes the Joint Committee on Structural Safety (JCSS) recommended procedure - CodeCal - for the practical implementation of reliability based code calibration of LRFD based design codes....

  14. Reduced Ambiguity Calibration for LOFAR

    CERN Document Server

    Yatawatta, Sarod

    2012-01-01

    Interferometric calibration always yields non unique solutions. It is therefore essential to remove these ambiguities before the solutions could be used in any further modeling of the sky, the instrument or propagation effects such as the ionosphere. We present a method for LOFAR calibration which does not yield a unitary ambiguity, especially under ionospheric distortions. We also present exact ambiguities we get in our solutions, in closed form. Casting this as an optimization problem, we also present conditions for this approach to work. The proposed method enables us to use the solutions obtained via calibration for further modeling of instrumental and propagation effects. We provide extensive simulation results on the performance of our method. Moreover, we also give cases where due to degeneracy, this method fails to perform as expected and in such cases, we suggest exploiting diversity in time, space and frequency.

  15. Astrid-2 EMMA Magnetic Calibration

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Risbo, Torben

    1998-01-01

    The Swedish micro-satellite Astrid-2 contains a tri-axial fluxgate magnetometer with the sensor co-located with a Technical University of Denmark (DTU) star camera for absolute attitude, and extended about 0.9 m on a hinged boom. The magnetometer is part of the RIT EMMA electric and magnetic fields...... Survey of Sweden near Stockholm on the night of May 15.-16., 1997. The magnetic calibration and the intercalibration between the star camera and the magnetic sensor was performed by measuring the Earth's magnetic field and simultaneously observing the star sky with the camera. The rotation matrix between...... fit calibration parameters. Owing to time shortage, we did not evaluate the temperature coefficients of the flight sensor calibration parameters. However, this was done for an identical flight spare magnetometer sensor at the magnetic coil facility belonging to the Technical University of Braunschweig...

  16. On chromatic and geometrical calibration

    DEFF Research Database (Denmark)

    Folm-Hansen, Jørgen

    1999-01-01

    of non-uniformity of the illumination of the image plane. Only the image deforming aberrations and the non-uniformity of illumination are included in the calibration models. The topics of the pinhole camera model and the extension to the Direct Linear Transform (DLT) are described. It is shown how......The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... the DLT can be extended with non-linear models of the common lens aberrations/errors some of them caused by manufacturing defects like decentering and thin prism distortion. The relation between a warping and the non-linear defects are shown. The issue of making a good resampling of an image by using...

  17. Comparison of different calibration approaches for chloramphenicol quantification in chicken muscle by ultra-high pressure liquid chromatography tandem mass spectrometry.

    Science.gov (United States)

    Pan, Xiao-Dong; Jiang, Wei; Wu, Ping-Gu

    2015-01-07

    Matrix-dependent signal suppression often occurs in quantitative analysis by ultra-high pressure liquid chromatography tandem mass spectrometry (UPLC-MS/MS). In this study, we investigated three calibration methods for compensation of signal suppression on chloramphenicol (CAP) quantification in chicken muscle. The data showed that the spiking recoveries by solvent standard calibration with a stable isotope labelled internal standard (SIL-IS) and matrix-matched standard calibration with a SIL-IS were significantly higher than by external matrix-matched standard calibration (P 0.05). The limit of detection (LOD) for external matrix matched standard calibration was 0.1 μg kg(-1), and that for SIL-IS calibration (including matrix matched and solvent dissolved standard) was 0.03 μg kg(-1).

  18. Evaluating uncertainties in the cross-calibration of parallel ion chambers used in electron beam radiotherapy; Avaliacao das incertezas no processo de calibracao cruzada de camaras de placas paralelas utilizadas em feixes de eletrons para radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Anderson, Ernani; Travassos, Paulo; Ferreira, Max da Silva; Carvalho, Samira Marques de; Silva, Michele Maria da; Peixoto, Jose Guilherme Pereira, E-mail: ernanianderson@hotmail.com [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil); Salmon Junior, Helio Augusto [Grupo COI - Clinicas Oncologicas, Rio de Janeiro, RJ (Brazil)

    2015-07-01

    This study aims to estimative the combined standard uncertainty for a detector parallel plate used for dosimetry of electron beams in linear accelerators for radiotherapy, which has been calibrated by the cross-calibration method. Keeping the combined standard uncertainty next of the uncertainty informed in the calibration certificate of the reference chamber, become possible establish the calibration factor of the detector. The combined standard uncertainty obtained in this study was 2.5 %. (author)

  19. Quantitative Analysis of Sulfide Minerals by Laser Ablation-Inductively Coupled Plasma-Mass Spectrometry Using Glass Reference Materials with Matrix Normalization Plus Sulfur Internal Standardization Calibration%玻璃标样结合硫内标归一定量技术在激光剥蚀-等离子体质谱分析硫化物矿物中的应用

    Institute of Scientific and Technical Information of China (English)

    袁继海; 詹秀春; 范晨子; 赵令浩; 孙冬阳; 贾泽荣; 胡明月; 蒯丽君

    2012-01-01

    A novel strategy for microanalysis of sulfides by laser ablation (LA)-ICP-MS was established. In this method, the relative sensitivity factor of Ca relative to S in anhydrite mineral reference material was taken as a transition bridge, by which the relative sensitivity factors of interesting elements in glass reference materials relative to Ca could be converted into relative sensitivity factors relative to S by the transition bridge, then the quantitative analysis of multi-elements in sulfide minerals by multi-glass reference materials with matrix normalization plus sulfur internal standardizationcalibration was carried out. 20 elements in the American polymetal sulfide mineral reference material MASS-1 were analyzed using this new method. The relative errors of major elements in MASS-1 were less than 10% ? And the results of trace elements with reference values were nearly within the uncertainty of the preliminary values. Multi-elements in 12 sulfide single minerals were analyzed by applying this new method. The relative errors of the greatest number of major elements were less than 10% , with which the results of most major elements were accurate than those obtained by MASS-1 as calibration standard with matrix normalization plus internal standardization or internal standard calibration. And the results to trace elements agreed well with the calibrated results by MASS-1 with matrix normalization plus internal standardization or internal standard calibration. This method overcomes the problem of non-matrix matched standards, enables to accurately determine the major composition of sulfur in sulfide minerals and suggests a new approach for analysis of sulfide minerals.%以硬石膏矿物标样中Ca相对于S的灵敏度因子为基准,将玻璃标样中主量和痕量元素相对于Ca的灵敏度因子转换成元素相对于S的灵敏度因子,建立了多玻璃标样结合硫内标归一定量技术分析硫化物单矿物多元素的新方法.利用本方法

  20. Instrument Calibration and Certification Procedure

    Energy Technology Data Exchange (ETDEWEB)

    Davis, R. Wesley [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2016-05-31

    The Amptec 640SL-2 is a 4-wire Kelvin failsafe resistance meter, designed to reliably use very low-test currents for its resistance measurements. The 640SL-1 is a 2-wire version, designed to support customers using the Reynolds Industries type 311 connector. For both versions, a passive (analog) dual function DC Milliameter/Voltmeter allows the user to verify the actual 640SL output current level and the open circuit voltage on the test leads. This procedure includes tests of essential performance parameters. Any malfunction noticed during calibration, whether specifically tested for or not, shall be corrected before calibration continues or is completed.

  1. 1 {\\Omega}-10 k{\\Omega} high precision transportable setup to calibrate multifunction electrical instruments

    CERN Document Server

    Capra, P P

    2015-01-01

    A temperature controlled 1 {\\Omega}-10 k{\\Omega} standard Resistors transportable setup was developed at National Institute of Metrological Research, (INRIM) for the calibration and adjustment of multifunction electrical instruments. The two Standards consist respectively of two 10 {\\Omega} and 100 k{\\Omega} parallel connected resistors nets inserted in a temperature controlled aluminium box. Novelty of the realization is the oil insertion of the 1 {\\Omega} net with its internal connectors lowering the thermo-electromotive forces (emfs) effects. Short and mid-term stabilities of the setup Standards resulted on the order and in some cases better than other top level 1 {\\Omega} and 10 k{\\Omega} commercial Standards. The transport effect turning off the setup temperature control did not cause appreciable measurement deviations on the two Standards. The Standards uncertainties meet those requested by DMMs and MFCs manufacturers to calibrate and adjust these instruments. A test to adjust a multifunction calibrator...

  2. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    Science.gov (United States)

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

  3. Calibration method for the reference parameter in Fernald and Klett inversion combining Raman and Elastic return

    Science.gov (United States)

    Ji, Hongzhu; Chen, Siying; Zhang, Yinchao; Chen, He; Guo, Pan; Chen, Hao

    2017-02-01

    A calibration method is proposed to invert the extinction coefficient for Fernald and Klett inversion by using the particle backscattering coefficient inversed with Raman and Elastic return signals. The calibration method is analyzed theoretically and experimentally, the inversion accuracy can be improved by removing the dependence on reference altitudes and intervals in conventional calibration methods, which resulted from the introduction of backscattering coefficient with relatively higher accuracy obtained by Raman-Mie inversion method. The standard deviation of this new calibration method can be reduced by about 20×, compared to that of the conventional calibration methods of Fernald and Klett inversion. And, the more stable effective inversed range with this new calibration method can be obtained by removing the dimple phenomenon in clouds position.

  4. 3C 286: a bright, compact, stable, and highly polarized calibrator for millimeter-wavelength observations

    CERN Document Server

    Agudo, Ivan; Wiesemeyer, Helmut; Molina, Sol N; Casadio, Carolina; Gomez, Jose L; Emmanoulopoulos, Dimitrios

    2012-01-01

    Context. A number of millimeter and submillimeter facilities with linear polarization observing capabilities have started operating during last years. These facilities, as well as other previous millimeter telescopes and interferometers, require bright and stable linear polarization calibrators to calibrate new instruments and to monitor their instrumental polarization. The current limited number of adequate calibrators implies difficulties in the acquisition of these calibration observations. Aims. Looking for additional linear polarization calibrators in the millimeter spectral range, in mid-2006 we started monitoring 3C 286, a standard and highly stable polarization calibrator for radio observations. Methods. Here we present the 3 and 1mm monitoring observations obtained between September 2006 and October 2011 with the XPOL polarimeter on the IRAM 30m Millimeter Telescope. Results. Our observations show that 3C 286 is a bright source of constant total flux with 3mm flux density S_3mm = (0.90 \\pm 0.02) Jy. ...

  5. Calibrating Data from the Hinode/X-Ray Telescope and Associated Uncertainties

    CERN Document Server

    Kobelski, Adam R; Weber, Mark A; McKenzie, David E; Reeves, Katharine K

    2013-01-01

    The X-Ray Telescope (XRT) onboard the Hinode satellite, launched 23 September 2006 by the Japanese Aerospace Exploration Agency (JAXA) is a joint mission between Japan, the United States, and the United Kingdom to study the solar corona. In particular XRT was designed to study solar plasmas with temperatures between 1 and 10 MK with $\\approx1''$ pixels ($\\approx2''$ resolution). Prior to analysis, the data product from this instrument must be properly calibrated and data values quantified in order to assess accurately the information contained within. We present here the standard methods of calibration for these data. The calibration is performed on an empirical basis which uses the least complicated correction that accurately describes the data while suppressing spurious features. By analyzing the uncertainties remaining in the data after calibration, we conclude that the procedure is successful, as the remaining uncertainty after calibration is dominated by photon noise. This calibration software is availab...

  6. Energy calibration of very inclined air showers

    Energy Technology Data Exchange (ETDEWEB)

    Dembinski, Hans [Institut fuer Experimentelle Kernphysik, Karlsruher Institut fuer Technologie (KIT), Karlsruhe (Germany); Roth, Markus [Institut fuer Kernphysik, Karlsruher Institut fuer Technologie (KIT), Karlsruhe (Germany); Hebbeker, Thomas [III. Physikalisches Institut A, RWTH Aachen (Germany)

    2010-07-01

    The Pierre Auger Observatory detects extensive air showers which are initiated by ultra-high energy cosmic rays. The properties of the cosmic rays are derived indirectly from the air shower observation. The surface detector of the observatory is well suited to detect showers with zenith angles from 0 to 90 . Standard analyses focus on so called vertical showers with inclinations smaller than 60 . Showers with larger zenith angles are called very inclined showers. Both have distinct experimental signatures which require separate event reconstructions. The ground signal of very inclined air showers is muon dominated and the energy reconstruction uses the variable R{sub {mu}} as an estimator for the cosmic ray energy which is proportional to the total number of muons N{sub {mu}} on the ground. The talk will focus on the energy calibration of R{sub {mu}} with events observed simultaneously in the surface and fluorescence detector of the observatory. This calibration procedure also offers the unique opportunity to derive the shower-to-shower fluctuations of R{sub {mu}} which are sensitive to the cosmic ray mass composition.

  7. A Generic Algorithm for IACT Optical Efficiency Calibration using Muons

    CERN Document Server

    Mitchell, A M W; Parsons, R D

    2015-01-01

    Muons produced in Extensive Air Showers (EAS) generate ring-like images in Imaging Atmospheric Cherenkov Telescopes when travelling near parallel to the optical axis. From geometrical parameters of these images, the absolute amount of light emitted may be calculated analytically. Comparing the amount of light recorded in these images to expectation is a well established technique for telescope optical efficiency calibration. However, this calculation is usually performed under the assumption of an approximately circular telescope mirror. The H.E.S.S. experiment entered its second phase in 2012, with the addition of a fifth telescope with a non-circular 600m$^2$ mirror. Due to the differing mirror shape of this telescope to the original four H.E.S.S. telescopes, adaptations to the standard muon calibration were required. We present a generalised muon calibration procedure, adaptable to telescopes of differing shapes and sizes, and demonstrate its performance on the H.E.S.S. II array.

  8. Calibration technology in application of robot-laser scanning system

    Science.gov (United States)

    Ren, YongJie; Yin, ShiBin; Zhu, JiGui

    2012-11-01

    A system composed of laser sensor and 6-DOF industrial robot is proposed to obtain complete three-dimensional (3-D) information of the object surface. Suitable for the different combining ways of laser sensor and robot, a new method to calibrate the position and pose between sensor and robot is presented. By using a standard sphere with known radius as a reference tool, the rotation and translation matrices between the laser sensor and robot are computed, respectively in two steps, so that many unstable factors introduced in conventional optimization methods can be avoided. The experimental results show that the accuracy of the proposed calibration method can be achieved up to 0.062 mm. The calibration method is also implemented into the automated robot scanning system to reconstruct a car door panel.

  9. Modeling of germanium detector and its sourceless calibration

    Directory of Open Access Journals (Sweden)

    Steljić Milijana

    2008-01-01

    Full Text Available The paper describes the procedure of adapting a coaxial high-precision germanium detector to a device with numerical calibration. The procedure includes the determination of detector dimensions and establishing the corresponding model of the system. In order to achieve a successful calibration of the system without the usage of standard sources, Monte Carlo simulations were performed to determine its efficiency and pulse-height response function. A detailed Monte Carlo model was developed using the MCNP-5.0 code. The obtained results have indicated that this method represents a valuable tool for the quantitative uncertainty analysis of radiation spectrometers and gamma-ray detector calibration, thus minimizing the need for the deployment of radioactive sources.

  10. Self-calibrating lateral scanning white-light interferometer

    Energy Technology Data Exchange (ETDEWEB)

    Munteanu, Florin

    2010-04-20

    Lateral scanning white-light interferometry represents an attractive alternative to the standard white-light interferometry. Its main advantage over the latter procedure consists in the ability to scan large samples continuously, without the need of a cumbersome stitching procedure. Presently, the main drawback in the path of large-scale industrial acceptance of this method is the need for careful calibration of the tilt angle prior to each measurement. A novel self-calibration approach is presented. Using the data acquired during the normal scanning process, the need of an initial tilt angle calibration is eliminated and on-the-fly system adjustments for the best signal-to-noise ratio can be performed without an increase in the measurement time dictated by recalibration.

  11. Calibration procedure for a laser triangulation scanner with uncertainty evaluation

    Science.gov (United States)

    Genta, Gianfranco; Minetola, Paolo; Barbato, Giulio

    2016-11-01

    Most of low cost 3D scanning devices that are nowadays available on the market are sold without a user calibration procedure to correct measurement errors related to changes in environmental conditions. In addition, there is no specific international standard defining a procedure to check the performance of a 3D scanner along time. This paper aims at detailing a thorough methodology to calibrate a 3D scanner and assess its measurement uncertainty. The proposed procedure is based on the use of a reference ball plate and applied to a triangulation laser scanner. Experimental results show that the metrological performance of the instrument can be greatly improved by the application of the calibration procedure that corrects systematic errors and reduces the device's measurement uncertainty.

  12. The New Approach to Camera Calibration - GCPs or TLS Data?

    Science.gov (United States)

    Markiewicz, J.; Podlasiak, P.; Kowalczyk, M.; Zawieska, D.

    2016-06-01

    Camera calibration is one of the basic photogrammetric tasks responsible for the quality of processed products. The majority of calibration is performed with a specially designed test field or during the self-calibration process. The research presented in this paper aims to answer the question of whether it is necessary to use control points designed in the standard way for determination of camera interior orientation parameters. Data from close-range laser scanning can be used as an alternative. The experiments shown in this work demonstrate the potential of laser measurements, since the number of points that may be involved in the calculation is much larger than that of commonly used ground control points. The problem which still exists is the correct and automatic identification of object details in the image, taken with a tested camera, as well as in the data set registered with the laser scanner.

  13. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-06-02

    This study addresses the effect of calibration methodologies on calibration responsivities and the resulting impact on radiometric measurements. The calibration responsivities used in this study are provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The BORCAL method provides outdoor calibration responsivity of pyranometers and pyrheliometers at a 45 degree solar zenith angle and responsivity as a function of solar zenith angle determined by clear-sky comparisons to reference irradiance. The BORCAL method also employs a thermal offset correction to the calibration responsivity of single-black thermopile detectors used in pyranometers. Indoor calibrations of radiometers by their manufacturers are performed using a stable artificial light source in a side-by-side comparison of the test radiometer under calibration to a reference radiometer of the same type. These different methods of calibration demonstrated 1percent to 2 percent differences in solar irradiance measurement. Analyzing these values will ultimately enable a reduction in radiometric measurement uncertainties and assist in developing consensus on a standard for calibration.

  14. Effective radiation attenuation calibration for breast density: compression thickness influences and correction

    Directory of Open Access Journals (Sweden)

    Thomas Jerry A

    2010-11-01

    Full Text Available Abstract Background Calibrating mammograms to produce a standardized breast density measurement for breast cancer risk analysis requires an accurate spatial measure of the compressed breast thickness. Thickness inaccuracies due to the nominal system readout value and compression paddle orientation induce unacceptable errors in the calibration. Method A thickness correction was developed and evaluated using a fully specified two-component surrogate breast model. A previously developed calibration approach based on effective radiation attenuation coefficient measurements was used in the analysis. Water and oil were used to construct phantoms to replicate the deformable properties of the breast. Phantoms consisting of measured proportions of water and oil were used to estimate calibration errors without correction, evaluate the thickness correction, and investigate the reproducibility of the various calibration representations under compression thickness variations. Results The average thickness uncertainty due to compression paddle warp was characterized to within 0.5 mm. The relative calibration error was reduced to 7% from 48-68% with the correction. The normalized effective radiation attenuation coefficient (planar representation was reproducible under intra-sample compression thickness variations compared with calibrated volume measures. Conclusion Incorporating this thickness correction into the rigid breast tissue equivalent calibration method should improve the calibration accuracy of mammograms for risk assessments using the reproducible planar calibration measure.

  15. Providing radiometric traceability for the calibration home base of DLR by PTB

    Energy Technology Data Exchange (ETDEWEB)

    Taubert, D. R.; Hollandt, J.; Sperfeld, P.; Pape, S.; Hoepe, A.; Hauer, K.-O. [Physikalisch-Technische Bundesanstalt, Braunschweig und Berlin, 10587 Berlin (Germany); Gege, P.; Schwarzmaier, T.; Lenhard, K.; Baumgartner, A. [Deutsches Zentrum fuer Luft- und Raumfahrt, Institut fuer Methodik der Fernerkundung, 82234 Oberpfaffenhofen (Germany)

    2013-05-10

    A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance E{sub {lambda}}({lambda}) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor {beta}{sub 0 Degree-Sign :45 Degree-Sign }({lambda}) of the RASTA reflection standard was calibrated in a 0 Degree-Sign :45 Degree-Sign -viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k= 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independent from procedure I and allowed to cover the entire spectral range of RASTA from 350 nm to 2500 nm. In the second procedure, the 0 Degree-Sign :45 Degree-Sign -viewing geometry spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign }({lambda}) of RASTA was directly calibrated at the Spectral Radiance Comparator Facility (SRCF) of PTB. The relative uncertainties for this calibration procedure range from 0.8 % in the visible up to 7.5 % at 2500 nm (k= 1). In the overlapping spectral range of both calibration procedures the calculated spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign ,calc}({lambda}) from procedure I is in good agreement with the direct measurement of procedure II, i.e. well within the combined expanded uncertainties (k= 2) of both procedures.

  16. Providing radiometric traceability for the calibration home base of DLR by PTB

    Science.gov (United States)

    Taubert, D. R.; Hollandt, J.; Sperfeld, P.; Pape, S.; Höpe, A.; Hauer, K.-O.; Gege, P.; Schwarzmaier, T.; Lenhard, K.; Baumgartner, A.

    2013-05-01

    A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance Eλ(λ) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor β0°:45°(λ) of the RASTA reflection standard was calibrated in a 0°:45°-viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k = 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independent from procedure I and allowed to cover the entire spectral range of RASTA from 350 nm to 2500 nm. In the second procedure, the 0°:45°-viewing geometry spectral radiance Lλ,0°:45°(λ) of RASTA was directly calibrated at the Spectral Radiance Comparator Facility (SRCF) of PTB. The relative uncertainties for this calibration procedure range from 0.8 % in the visible up to 7.5 % at 2500 nm (k = 1). In the overlapping spectral range of both calibration procedures the calculated spectral radiance Lλ,0°:45°,calc(λ) from procedure I is in good agreement with the direct measurement of procedure II, i.e. well within the combined expanded uncertainties (k = 2) of both procedures.

  17. High-precision efficiency calibration of a high-purity co-axial germanium detector

    Energy Technology Data Exchange (ETDEWEB)

    Blank, B., E-mail: blank@cenbg.in2p3.fr [Centre d' Etudes Nucléaires de Bordeaux Gradignan, UMR 5797, CNRS/IN2P3, Université de Bordeaux, Chemin du Solarium, BP 120, 33175 Gradignan Cedex (France); Souin, J.; Ascher, P.; Audirac, L.; Canchel, G.; Gerbaux, M.; Grévy, S.; Giovinazzo, J.; Guérin, H.; Nieto, T. Kurtukian; Matea, I. [Centre d' Etudes Nucléaires de Bordeaux Gradignan, UMR 5797, CNRS/IN2P3, Université de Bordeaux, Chemin du Solarium, BP 120, 33175 Gradignan Cedex (France); Bouzomita, H.; Delahaye, P.; Grinyer, G.F.; Thomas, J.C. [Grand Accélérateur National d' Ions Lourds, CEA/DSM, CNRS/IN2P3, Bvd Henri Becquerel, BP 55027, F-14076 CAEN Cedex 5 (France)

    2015-03-11

    A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and γ-ray source measurements have been compared to Monte-Carlo simulations to adjust the dimensions of a detector model. For this purpose, standard calibration sources and short-lived online sources have been used. The resulting efficiency calibration reaches the precision needed e.g. for branching ratio measurements of super-allowed β decays for tests of the weak-interaction standard model.

  18. Inter-laboratory project q calibration of SANS instruments using silver behenate

    Energy Technology Data Exchange (ETDEWEB)

    Ikram, Abarrul; Gunawan; Edy Giri, Putra [Indonesia National Nuclear Energy Agency (BATAN) (Indonesia); Suzuki, Jun-ichi [Japan Atomic Energy Research Inst., Tokyo (Japan); Knott, Robert [Australian Nuclear Science and Technology Organisation (ANSTO) (Australia)

    2000-10-01

    The inter-laboratory project for q-calibration of SANS (small angle neutron scattering) using silver behenate was carried out among Indonesia National Nuclear Energy Agency (BATAN), Japan Atomic Energy Research Institute (JAERI) and Australian Nuclear Science and Technology Organization (ANSTO). The standard sample of silver behenate, [CH{sub 3}(CH{sub 2}){sub 20}COOAg](AgBE), has been assessed as an international standard for the calibration of both x-ray and neutron scattering instruments. The results indicate excellent agreement for q calibration obtained among the three laboratories, BATAN, JAERI and ANSTO. (Y. Kazumata)

  19. Scalar Calibration of Vector Magnetometers

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz;

    2000-01-01

    The calibration parameters of a vector magnetometer are estimated only by the use of a scalar reference magnetometer. The method presented in this paper differs from those previously reported in its linearized parametrization. This allows the determination of three offsets or signals in the absence...

  20. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Villanueva, Héctor

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this repor...

  1. Practical intraoperative stereo camera calibration.

    Science.gov (United States)

    Pratt, Philip; Bergeles, Christos; Darzi, Ara; Yang, Guang-Zhong

    2014-01-01

    Many of the currently available stereo endoscopes employed during minimally invasive surgical procedures have shallow depths of field. Consequently, focus settings are adjusted from time to time in order to achieve the best view of the operative workspace. Invalidating any prior calibration procedure, this presents a significant problem for image guidance applications as they typically rely on the calibrated camera parameters for a variety of geometric tasks, including triangulation, registration and scene reconstruction. While recalibration can be performed intraoperatively, this invariably results in a major disruption to workflow, and can be seen to represent a genuine barrier to the widespread adoption of image guidance technologies. The novel solution described herein constructs a model of the stereo endoscope across the continuum of focus settings, thereby reducing the number of degrees of freedom to one, such that a single view of reference geometry will determine the calibration uniquely. No special hardware or access to proprietary interfaces is required, and the method is ready for evaluation during human cases. A thorough quantitative analysis indicates that the resulting intrinsic and extrinsic parameters lead to calibrations as accurate as those derived from multiple pattern views.

  2. CALIBRATION OF THE INFRARED OPTOMETER

    Science.gov (United States)

    An infrared optometer for measuring the absolute status of accommodation is subject to a constant error not associated with chromatic aberration or...on optometer accuracy as long as the pupil does not vignette the optometer beam. A modification is described for calibrating the infrared optometer ...for an individual subject without using trial lenses or a subjective optometer . (Author)

  3. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Kock, Carsten Weber; Vesth, Allan

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  4. Measurement System and Calibration report

    DEFF Research Database (Denmark)

    Vesth, Allan; Kock, Carsten Weber

    The report describes power curve measurements carried out on a given wind turbine. The measurements are carried out in accordance to Ref. [1]. A site calibration has been carried out; see Ref. [2], and the measured flow correction factors for different wind directions are used in the present...

  5. Networks of Absolute Calibration Stars for SST, AKARI, and WISE

    Science.gov (United States)

    Cohen, M.

    2007-04-01

    I describe the Cohen-Walker-Witteborn (CWW) network of absolute calibration stars built to support ground-based, airborne, and space-based sensors, and how they are used to calibrate instruments on the SPITZER Space Telescope (SST and Japan's AKARI (formerly ASTRO-F), and to support NASA's planned MidEx WISE (the Wide-field Infrared Survey Explorer). All missions using this common calibration share a self-consistent framework embracing photometry and low-resolution spectroscopy. CWW also underpins COBE/DIRBE several instruments used on the Kuiper Airborne Observatory ({KAO}), the joint Japan-USA ``IR Telescope in Space" (IRTS) Near-IR and Mid-IR spectrometers, the European Space Agency's IR Space Observatory (ISO), and the US Department of Defense's Midcourse Space eXperiment (MSX). This calibration now spans the far-UV to mid-infrared range with Sirius (one specific Kurucz synthetic spectrum) as basis, and zero magnitude defined from another Kurucz spectrum intended to represent an ideal Vega (not the actual star with its pole-on orientation and mid-infrared dust excess emission). Precision 4-29 μm radiometric measurements on MSX validate CWW's absolute Kurucz spectrum of Sirius, the primary, and a set of bright K/MIII secondary standards. Sirius is measured to be 1.0% higher than predicted. CWW's definitions of IR zero magnitudes lie within 1.1% absolute of MSX measurements. The US Air Force Research Laboratory's independent analysis of on-orbit {MSX} stellar observations compared with emissive reference spheres show CWW primary and empirical secondary spectra lie well within the ±1.45% absolute uncertainty associated with this 15-year effort. Our associated absolute calibration for the InfraRed Array Camera (IRAC) on the SST lies within ˜2% of the recent extension of the calibration of the Hubble Space Telescope's STIS instrument to NICMOS (Bohlin, these Proceedings), showing the closeness of these two independent approaches to calibration.

  6. Calibration methodology and performance characterization of a polarimetric hyperspectral imager

    Science.gov (United States)

    Holder, Joel G.; Martin, Jacob A.; Pitz, Jeremey; Pezzaniti, Joseph L.; Gross, Kevin C.

    2014-05-01

    Polarimetric hyperspectral imaging (P-HSI) has the potential to improve target detection, material identification, and background characterization over conventional hyperspectral imaging and polarimetric imaging. To fully exploit the spectro-polarimetric signatures captured by such an instrument, a careful calibration process is required to remove the spectrally- and polarimetrically-dependent system response (gain). Calibration of instruments operating in the long-wave infrared (LWIR, 8μm to 12 μm) is further complicated by the polarized spectral radiation generated within the instrument (offset). This paper presents a calibration methodology developed for a LWIR Telops Hyper-Cam modified for polarimetry by replacing the entrance window with a rotatable holographic wire-grid polarizer (4000 line/mm, ZnSe substrate, 350:1 extinction ratio). A standard Fourier-transform spectrometer (FTS) spectro-radiometric calibration is modified to include a Mueller-matrix approach to account for polarized transmission through and polarized selfemission from each optical interface. It is demonstrated that under the ideal polarizer assumption, two distinct blackbody measurements at polarizer angles of 0°, 45°, 90°, and 135° are sufficient to calibrate the system for apparent degree-of-linear-polarization (DoLP) measurements. Noise-equivalent s1, s2, and DoLP are quantified using a wide-area blackbody. A polarization-state generator is used to determine the Mueller deviation matrix. Finally, a realistic scene involving buildings, cars, sky radiance, and natural vegetation is presented.

  7. Research on the calibration of ultraviolet energy meters

    Science.gov (United States)

    Lin, Fangsheng; Yin, Dejin; Li, Tiecheng; Lai, Lei; Xia, Ming

    2016-10-01

    Ultraviolet (UV) radiation is a kind of non-lighting radiation with the wavelength range from 100nm to 400nm. Ultraviolet irradiance meters are now widely used in many areas. However, as the development of science and technology, especially in the field of light-curing industry, there are more and more UV energy meters or UV-integrators need to be measured. Because the structure, wavelength band and measured power intensity of UV energy meters are different from traditional UV irradiance meters, it is important for us to take research on the calibration. With reference to JJG879-2002, we SIMT have independently developed the UV energy calibration device and the standard of operation and experimental methods for UV energy calibration in detail. In the calibration process of UV energy meter, many influencing factors will affect the final results, including different UVA-band UV light sources, different spectral response for different brands of UV energy meters, instability and no uniformity of UV light source and temperature. Therefore we need to take all of these factors into consideration to improve accuracy in UV energy calibration.

  8. Precision Spectrophotometric Calibration System for Dark Energy Instruments

    Energy Technology Data Exchange (ETDEWEB)

    Schubnell, Michael S.

    2015-06-30

    For this research we build a precision calibration system and carried out measurements to demonstrate the precision that can be achieved with a high precision spectrometric calibration system. It was shown that the system is capable of providing a complete spectrophotometric calibration at the sub-pixel level. The calibration system uses a fast, high precision monochromator that can quickly and efficiently scan over an instrument’s entire spectral range with a spectral line width of less than 0.01 nm corresponding to a fraction of a pixel on the CCD. The system was extensively evaluated in the laboratory. Our research showed that a complete spectrophotometric calibration standard for spectroscopic survey instruments such as DESI is possible. The monochromator precision and repeatability to a small fraction of the DESI spectrograph LSF was demonstrated with re-initialization on every scan and thermal drift compensation by locking to multiple external line sources. A projector system that mimics telescope aperture for point source at infinity was demonstrated.

  9. The DICE calibration project: design, characterization, and first results

    CERN Document Server

    Regnault, N; Schahmanèche, K; Guillou, L Le; Antilogus, P; Astier, P; Barrelet, E; Betoule, M; Bongard, S; Cuillandre, J -C; Juramy, C; Pain, R; Rocci, P -F; Tisserand, P; Villa, F

    2015-01-01

    We describe the design, operation, and first results of a photometric calibration project, called DICE (Direct Illumination Calibration Experiment), aiming at achieving precise instrumental calibration of optical telescopes. The heart of DICE is an illumination device composed of 24 narrow-spectrum, high-intensity, light-emitting diodes (LED) chosen to cover the ultraviolet-to-near-infrared spectral range. It implements a point-like source placed at a finite distance from the telescope entrance pupil, yielding a flat field illumination that covers the entire field of view of the imager. The purpose of this system is to perform a lightweight routine monitoring of the imager passbands with a precision better than 5 per-mil on the relative passband normalisations and about 3{\\AA} on the filter cutoff positions. The light source is calibrated on a spectrophotometric bench. As our fundamental metrology standard, we use a photodiode calibrated at NIST. The radiant intensity of each beam is mapped, and spectra are m...

  10. Calibration of Instruments for Measuring Wind Velocity and Direction

    Science.gov (United States)

    Vogler, Raymond D.; Pilny, Miroslav J.

    1950-01-01

    Signal Corps wind equipment AN/GMQ-1 consisting of a 3-cup anemometer and wind vane was calibrated for wind velocities from 1 to 200 miles per hour. Cup-shaft failure prevented calibration at higher wind velocities. The action of the wind vane was checked and found to have very poor directional accuracy below a velocity of 8 miles per hour. After shaft failure was reported to the Signal Corps, the cup rotors were redesigned by strengthening the shafts for better operation at high velocities. The anemometer with the redesigned cup rotors was recalibrated, but cup-shaft failure occurred again at a wind velocity of approximately 220 miles per hour. In the course of this calibration two standard generators were checked for signal output variation, and a wind-speed meter was calibrated for use with each of the redesigned cup rotors. The variation of pressure coefficient with air-flow direction at four orifices on a disk-shaped pitot head was obtained for wind velocities of 37.79 53.6, and 98.9 miles per hour. A pitot-static tube mounted in the nose of a vane was calibrated up to a dynamic pressure of 155 pounds per square foot, or approximately 256 miles per hour,

  11. Photometric Calibrations of Gemini Images of NGC 6253

    Science.gov (United States)

    Pearce, Sean; Jeffery, Elizabeth

    2017-01-01

    We present preliminary results of our analysis of the metal-rich open cluster NGC 6253 using imaging data from GMOS on the Gemini-South Observatory. These data are part of a larger project to observe the effects of high metallicity on white dwarf cooling processes, especially the white dwarf cooling age, which have important implications on the processes of stellar evolution. To standardize the Gemini photometry, we have also secured imaging data of both the cluster and standard star fields using the 0.6-m SARA Observatory at CTIO. By analyzing and comparing the standard star fields of both the SARA data and the published Gemini zero-points of the standard star fields, we will calibrate the data obtained for the cluster. These calibrations are an important part of the project to obtain a standardized deep color-magnitude diagram to analyze the cluster. We present the process of verifying our standardization process. With a standardized CMD, we also present an analysis of the cluster's main sequence turn off age.

  12. Fluorescence of ceramic color standards.

    Science.gov (United States)

    Koo, Annette; Clare, John F; Nield, Kathryn M; Deadman, Andrew; Usadi, Eric

    2010-04-20

    Fluorescence has been found in color standards available for use in calibration and verification of color measuring instruments. The fluorescence is excited at wavelengths below about 600?nm and emitted above 700?nm, within the response range of silicon photodiodes, but at the edge of the response of most photomultipliers and outside the range commonly scanned in commercial colorimeters. The degree of fluorescence on two of a set of 12 glossy ceramic tiles is enough to introduce significant error when those tiles have been calibrated in one mode of measurement and are used in another. We report the nature of the fluorescence and the implications for color measurement.

  13. Radiation calibration for LWIR Hyperspectral Imager Spectrometer

    Science.gov (United States)

    Yang, Zhixiong; Yu, Chunchao; Zheng, Wei-jian; Lei, Zhenggang; Yan, Min; Yuan, Xiaochun; Zhang, Peizhong

    2014-11-01

    The radiometric calibration of LWIR Hyperspectral imager Spectrometer is presented. The lab has been developed to LWIR Interferometric Hyperspectral imager Spectrometer Prototype(CHIPED-I) to study Lab Radiation Calibration, Two-point linear calibration is carried out for the spectrometer by using blackbody respectively. Firstly, calibration measured relative intensity is converted to the absolute radiation lightness of the object. Then, radiation lightness of the object is is converted the brightness temperature spectrum by the method of brightness temperature. The result indicated †that this method of Radiation Calibration calibration was very good.

  14. Radio Interferometric Calibration Using a Riemannian Manifold

    CERN Document Server

    Yatawatta, Sarod

    2013-01-01

    In order to cope with the increased data volumes generated by modern radio interferometers such as LOFAR (Low Frequency Array) or SKA (Square Kilometre Array), fast and efficient calibration algorithms are essential. Traditional radio interferometric calibration is performed using nonlinear optimization techniques such as the Levenberg-Marquardt algorithm in Euclidean space. In this paper, we reformulate radio interferometric calibration as a nonlinear optimization problem on a Riemannian manifold. The reformulated calibration problem is solved using the Riemannian trust-region method. We show that calibration on a Riemannian manifold has faster convergence with reduced computational cost compared to conventional calibration in Euclidean space.

  15. SDSS-IV/MaNGA: Spectrophotometric Calibration Technique

    Science.gov (United States)

    Yan, Renbin; Tremonti, Christy; Bershady, Matthew A.; Law, David R.; Schlegel, David J.; Bundy, Kevin; Drory, Niv; MacDonald, Nicholas; Bizyaev, Dmitry; Blanc, Guillermo A.; Blanton, Michael R.; Cherinka, Brian; Eigenbrot, Arthur; Gunn, James E.; Harding, Paul; Hogg, David W.; Sánchez-Gallego, José R.; Sánchez, Sebastian F.; Wake, David A.; Weijmans, Anne-Marie; Xiao, Ting; Zhang, Kai

    2016-01-01

    Mapping Nearby Galaxies at Apache Point Observatory (MaNGA), one of three core programs in the Sloan Digital Sky Survey-IV, is an integral-field spectroscopic survey of roughly 10,000 nearby galaxies. It employs dithered observations using 17 hexagonal bundles of 2″ fibers to obtain resolved spectroscopy over a wide wavelength range of 3600-10300 Å. To map the internal variations within each galaxy, we need to perform accurate spectral surface photometry, which is to calibrate the specific intensity at every spatial location sampled by each individual aperture element of the integral field unit. The calibration must correct only for the flux loss due to atmospheric throughput and the instrument response, but not for losses due to the finite geometry of the fiber aperture. This requires the use of standard star measurements to strictly separate these two flux loss factors (throughput versus geometry), a difficult challenge with standard single-fiber spectroscopy techniques due to various practical limitations. Therefore, we developed a technique for spectral surface photometry using multiple small fiber-bundles targeting standard stars simultaneously with galaxy observations. We discuss the principles of our approach and how they compare to previous efforts, and we demonstrate the precision and accuracy achieved. MaNGA's relative calibration between the wavelengths of Hα and Hβ has an rms of 1.7%, while that between [N ii] λ6583 and [O ii] λ3727 has an rms of 4.7%. Using extinction-corrected star formation rates and gas-phase metallicities as an illustration, this level of precision guarantees that flux calibration errors will be sub-dominant when estimating these quantities. The absolute calibration is better than 5% for more than 89% of MaNGA's wavelength range.

  16. SDSS-IV/MaNGA: SPECTROPHOTOMETRIC CALIBRATION TECHNIQUE

    Energy Technology Data Exchange (ETDEWEB)

    Yan, Renbin; Sánchez-Gallego, José R. [Department of Physics and Astronomy, University of Kentucky, 505 Rose St., Lexington, KY 40506-0057 (United States); Tremonti, Christy; Bershady, Matthew A.; Eigenbrot, Arthur; Wake, David A. [Department of Astronomy, University of Winsconsin-Madison, 475 N. Charter Street, Madison, WI 53706-1582 (United States); Law, David R. [Space Telescope Science Institute, 3700 San Martin Drive, Baltimore, MD 21218 (United States); Schlegel, David J. [Physics Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720-8160 (United States); Bundy, Kevin [Kavli IPMU (WPI), UTIAS, The University of Tokyo, Kashiwa, Chiba 277-8583 (Japan); Drory, Niv [McDonald Observatory, Department of Astronomy, University of Texas at Austin, 1 University Station, Austin, TX 78712-0259 (United States); MacDonald, Nicholas [Department of Astronomy, Box 351580, University of Washington, Seattle, WA 98195 (United States); Bizyaev, Dmitry [Apache Point Observatory, P.O. Box 59, sunspot, NM 88349 (United States); Blanc, Guillermo A. [Departamento de Astronomía, Universidad de Chile, Camino el Observatorio 1515, Las Condes, Santiago (Chile); Blanton, Michael R.; Hogg, David W. [Center for Cosmology and Particle Physics, Department of Physics, New York University, 4 Washington Place, New York, NY 10003 (United States); Cherinka, Brian [Dunlap Institute for Astronomy and Astrophysics, University of Toronto, 50 St. George Street, Toronto, Ontario M5S 3H4 (Canada); Gunn, James E. [Department of Astrophysical Sciences, Princeton University, Princeton, NJ 08544 (United States); Harding, Paul [Department of Astronomy, Case Western Reserve University, Cleveland, OH 44106 (United States); Sánchez, Sebastian F., E-mail: yanrenbin@uky.edu [Instituto de Astronomia, Universidad Nacional Autonoma de Mexico, A.P. 70-264, 04510 Mexico D.F. (Mexico); and others

    2016-01-15

    Mapping Nearby Galaxies at Apache Point Observatory (MaNGA), one of three core programs in the Sloan Digital Sky Survey-IV, is an integral-field spectroscopic survey of roughly 10,000 nearby galaxies. It employs dithered observations using 17 hexagonal bundles of 2″ fibers to obtain resolved spectroscopy over a wide wavelength range of 3600–10300 Å. To map the internal variations within each galaxy, we need to perform accurate spectral surface photometry, which is to calibrate the specific intensity at every spatial location sampled by each individual aperture element of the integral field unit. The calibration must correct only for the flux loss due to atmospheric throughput and the instrument response, but not for losses due to the finite geometry of the fiber aperture. This requires the use of standard star measurements to strictly separate these two flux loss factors (throughput versus geometry), a difficult challenge with standard single-fiber spectroscopy techniques due to various practical limitations. Therefore, we developed a technique for spectral surface photometry using multiple small fiber-bundles targeting standard stars simultaneously with galaxy observations. We discuss the principles of our approach and how they compare to previous efforts, and we demonstrate the precision and accuracy achieved. MaNGA's relative calibration between the wavelengths of Hα and Hβ has an rms of 1.7%, while that between [N ii] λ6583 and [O ii] λ3727 has an rms of 4.7%. Using extinction-corrected star formation rates and gas-phase metallicities as an illustration, this level of precision guarantees that flux calibration errors will be sub-dominant when estimating these quantities. The absolute calibration is better than 5% for more than 89% of MaNGA's wavelength range.

  17. Primary Standards Laboratory report, 2nd half 1993

    Energy Technology Data Exchange (ETDEWEB)

    1994-05-01

    The Primary Standards Laboratory (PSL) operates a system-wide primary standards and calibration program for the US Department of Energy, Albuquerque Field Office (DOE/AL). The PSL mission is as follows: to develop and maintain primary standards; to calibrate electrical, physical, and radiation reference standards for customer laboratories (DOE/AL nuclear weapon contractors); to conduct the technical surveys and measurement audits of these laboratories; and to recommend and implement system-wide improvements. This report summarizes activities of the PSL for the second half of 1993 and provides information pertinent to the operation of the DOE/AL Standards and Calibration Program. Specific areas covered include development projects, improvement projects, calibration and special measurements, surveys and audits, customer service, and significant events. Appendixes include certifications and reports;; a discussion about commercial calibration laboratories; PSL memoranda (PSLM); test numbers from the National Institute of Standards and Technology (NIST), formerly the National Bureau of Standards (NBS); and DOE/PSL memoranda on the Standards and Calibration Program with emphasis on traceability of PSL calibrations.

  18. Calibration setting numbers for dose calibrators for the PET isotopes (52)Mn, (64)Cu, (76)Br, (86)Y, (89)Zr, (124)I.

    Science.gov (United States)

    Wooten, A Lake; Lewis, Benjamin C; Szatkowski, Daniel J; Sultan, Deborah H; Abdin, Kinda I; Voller, Thomas F; Liu, Yongjian; Lapi, Suzanne E

    2016-07-01

    For PET radionuclides, the radioactivity of a sample can be conveniently measured by a dose calibrator. These devices depend on a "calibration setting number", but many recommended settings from manuals were interpolated based on standard sources of other radionuclide(s). We conducted HPGe gamma-ray spectroscopy, resulting in a reference for determining settings in two types of vessels containing one of several PET radionuclides. Our results reiterate the notion that in-house, experimental calibrations are recommended for different radionuclides and vessels.

  19. SMAP RADAR Calibration and Validation

    Science.gov (United States)

    West, R. D.; Jaruwatanadilok, S.; Chaubel, M. J.; Spencer, M.; Chan, S. F.; Chen, C. W.; Fore, A.

    2015-12-01

    The Soil Moisture Active Passive (SMAP) mission launched on Jan 31, 2015. The mission employs L-band radar and radiometer measurements to estimate soil moisture with 4% volumetric accuracy at a resolution of 10 km, and freeze-thaw state at a resolution of 1-3 km. Immediately following launch, there was a three month instrument checkout period, followed by six months of level 1 (L1) calibration and validation. In this presentation, we will discuss the calibration and validation activities and results for the L1 radar data. Early SMAP radar data were used to check commanded timing parameters, and to work out issues in the low- and high-resolution radar processors. From April 3-13 the radar collected receive only mode data to conduct a survey of RFI sources. Analysis of the RFI environment led to a preferred operating frequency. The RFI survey data were also used to validate noise subtraction and scaling operations in the radar processors. Normal radar operations resumed on April 13. All radar data were examined closely for image quality and calibration issues which led to improvements in the radar data products for the beta release at the end of July. Radar data were used to determine and correct for small biases in the reported spacecraft attitude. Geo-location was validated against coastline positions and the known positions of corner reflectors. Residual errors at the time of the beta release are about 350 m. Intra-swath biases in the high-resolution backscatter images are reduced to less than 0.3 dB for all polarizations. Radiometric cross-calibration with Aquarius was performed using areas of the Amazon rain forest. Cross-calibration was also examined using ocean data from the low-resolution processor and comparing with the Aquarius wind model function. Using all a-priori calibration constants provided good results with co-polarized measurements matching to better than 1 dB, and cross-polarized measurements matching to about 1 dB in the beta release. During the

  20. QA experience at the University of Wisconsin accredited dosimetry calibration laboratory

    Energy Technology Data Exchange (ETDEWEB)

    DeWard, L.A.; Micka, J.A. [Univ. of Wisconsin, Madison, WI (United States)

    1993-12-31

    The University of Wisconsin Accredited Dosimetry Calibration Laboratory (UW ADCL) employs procedure manuals as part of its Quality Assurance (QA) program. One of these manuals covers the QA procedures and results for all of the UW ADCL measurement equipment. The QA procedures are divided into two main areas: QA for laboratory equipment and QA for external chambers sent for calibration. All internal laboratory equipment is checked and recalibrated on an annual basis, after establishing its consistency on a 6-month basis. QA for external instruments involves checking past calibration history as well as comparing to a range of calibration values for specific instrument models. Generally, the authors find that a chamber will have a variation of less than 0.5 % from previous Co-60 calibration factors, and falls within two standard deviations of previous calibrations. If x-ray calibrations are also performed, the energy response of the chamber is plotted and compared to previous instruments of the same model. These procedures give the authors confidence in the transfer of calibration values from National Institute of Standards and Technology (NIST).

  1. Shortwave Radiometer Calibration Methods Comparison and Resulting Solar Irradiance Measurement Differences: A User Perspective

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-11-21

    different methods of calibration demonstrated +1% to +2% differences in solar irradiance measurement. Analyzing these differences will ultimately help determine the uncertainty of the field radiometer data and guide the development of a consensus standard for calibration. Further advancing procedures for precisely calibrating radiometers to world reference standards that reduce measurement uncertainty will allow more accurate prediction of solar output and improve the bankability of solar projects.

  2. [In-Flight Radiometric Calibration for ZY-3 Satellite Multispectral Sensor by Modified Reflectance-Based Method].

    Science.gov (United States)

    Han, Jie; Xie, Yong; Gu, Xing-fa; Yu, Tao; Liu, Qi-yue; Gao, Rong-jun

    2015-03-01

    Through integrating multi-spectral sensor characteristics of ZY-3 satellite, a modified reflectance-based method is proposed and used to achieve ZY-3 satellite multispectral sensor in-flight radiometric calibration. This method chooses level 1A image as data source and establishes geometric model to get an accurate observation geometric parameters at calibration site according to the information provided in image auxiliary documentation, which can reduce the influences on the calibration accuracy from image resampling and observation geometry errors. We use two-point and multi-points methods to calculate the absolute radiometric calibration coefficients of ZY-3 satellite multispectral sensor based on the large campaign at Dongying city, Shan Dong province. Compared with ZY-3 official calibration coefficients, multi-points method has higher accuracy than two-point method. Through analyzing the dispersion between each calibration point and the fitting line, we find that the residual error of water calibration site is larger than others, which of green band is approximately 67.39%. Treating water calibration site as an error, we filter it out using 95.4% confidence level as standard and recalculate the calibration coefficients with multi-points method. The final calibration coefficients show that the relative differences of the first three bands are less than 2% and the last band is less than 5%, which manifests that the proposed radiometric calibration method can obtain accurate and reliable calibration coefficients and is useful for other similar satellites in future.

  3. Crop physiology calibration in the CLM

    Directory of Open Access Journals (Sweden)

    I. Bilionis

    2015-04-01

    scalable and adaptive scheme based on sequential Monte Carlo (SMC. The model showed significant improvement of crop productivity with the new calibrated parameters. We demonstrate that the calibrated parameters are applicable across alternative years and different sites.

  4. Calibration and Validation of Measurement System

    DEFF Research Database (Denmark)

    Kofoed, Jens Peter; Riemann, Sven; Knapp, Wilfried

    The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype.......The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype....

  5. A Sub-millikelvin Calibration Facility in the Range 0 °C to 30 °C

    Science.gov (United States)

    Bosma, R.; Peruzzi, A.; Van Breugel, R.; Bruin-Barendregt, C.

    2017-03-01

    A new sub-millikelvin calibration facility for the range 0 °C to 30 °C is described, that allows calibration of customer thermometers, other than standard platinum resistance thermometers, with an uncertainty lower than 1 millikelvin. The improvements with respect to the traditional calibration facility are reported with particular emphasis on the temperature control (better than 0.2 mK), resistance measurement and calibration procedure. The new facility was validated by using 6 standard platinum resistance thermometers and the calibration uncertainty in the range from 0 °C to 30 °C amounted to 0.31 mK-0.35 mK. To demonstrate the potentiality of this facility, two oceanographic thermometers, Sea-Bird Electronics SBE 3 and SBE 35, were calibrated with an expanded uncertainty of 0.8 mK (k=2).

  6. Application of the SKYRAD Improved Langley plot method for the in situ calibration of CIMEL Sun-sky photometers.

    Science.gov (United States)

    Campanelli, Monica; Estellés, Víctor; Tomasi, Claudio; Nakajima, Teruyuki; Malvestuto, Vincenzo; Martínez-Lozano, José Antonio

    2007-05-10

    The in situ procedure for determining the solar calibration constants, originally developed for the PREDE Sun-sky radiometers and based on a modified version of the Langley plot, was applied to a CIMEL instrument located in Valencia, Spain, not integrated into AERONET. Taking into account the different mechanical and electronic characteristics of the two radiometers, the method was adapted to the characteristics of the CIMEL instrument. The iterative procedure for the determination of the solar calibration constants was applied to a 3-year data set. The results were compared with the two sets of experimental calibration constants determined during this period using the standard Langley plot method. The agreement was found to be consistent with the experimental errors, and the method can definitely also be used to determine the solar calibration constant for the CIMEL instrument, improving its calibration. The method can be used provided the radiometer is previously calibrated for diffuse radiance using a standard lamp.

  7. Optimal Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard; Kroon, I. B.; Faber, M. H.

    1994-01-01

    Calibration of partial safety factors is considered in general, including classes of structures where no code exists beforehand. The partial safety factors are determined such that the difference between the reliability for the different structures in the class considered and a target reliability...... level is minimized. Code calibration on a decision theoretical basis is also considered and it is shown how target reliability indices can be calibrated. Results from code calibration for rubble mound breakwater designs are shown....

  8. Calibration of a Parallel Kinematic Machine Tool

    Institute of Scientific and Technical Information of China (English)

    HE Xiao-mei; DING Hong-sheng; FU Tie; XIE Dian-huang; XU Jin-zhong; LI Hua-feng; LIU Hui-lin

    2006-01-01

    A calibration method is presented to enhance the static accuracy of a parallel kinematic machine tool by using a coordinate measuring machine and a laser tracker. According to the established calibration model and the calibration experiment, the factual 42 kinematic parameters of BKX-I parallel kinematic machine tool are obtained. By circular tests the comparison is made between the calibrated and the uncalibrated parameters and shows that there is 80% improvement in accuracy of this machine tool.

  9. A Careful Consideration of the Calibration Concept

    Science.gov (United States)

    Phillips, S. D.; Estler, W. T.; Doiron, T.; Eberhardt, K. R.; Levenson, M. S.

    2001-01-01

    This paper presents a detailed discussion of the technical aspects of the calibration process with emphasis on the definition of the measurand, the conditions under which the calibration results are valid, and the subsequent use of the calibration results in measurement uncertainty statements. The concepts of measurement uncertainty, error, systematic error, and reproducibility are also addressed as they pertain to the calibration process. PMID:27500027

  10. Calibrating thermal behavior of electronics

    Energy Technology Data Exchange (ETDEWEB)

    Chainer, Timothy J.; Parida, Pritish R.; Schultz, Mark D.

    2017-01-03

    A method includes determining a relationship between indirect thermal data for a processor and a measured temperature associated with the processor, during a calibration process, obtaining the indirect thermal data for the processor during actual operation of the processor, and determining an actual significant temperature associated with the processor during the actual operation using the indirect thermal data for the processor during actual operation of the processor and the relationship.

  11. Nonlinear Observers for Gyro Calibration

    Science.gov (United States)

    Thienel, Julie; Sanner, Robert M.

    2003-01-01

    Nonlinear observers for gyro calibration are presented. The first observer estimates a constant gyro bias. The second observer estimates scale factor errors. The third observer estimates the gyro alignment for three orthogonal gyros. The convergence properties of all three observers are discussed. Additionally, all three observers are coupled with a nonlinear control algorithm. The stability of each of the resulting closed loop systems is analyzed. Simulated test results are presented for each system.

  12. Calibrating thermal behavior of electronics

    Science.gov (United States)

    Chainer, Timothy J.; Parida, Pritish R.; Schultz, Mark D.

    2016-05-31

    A method includes determining a relationship between indirect thermal data for a processor and a measured temperature associated with the processor, during a calibration process, obtaining the indirect thermal data for the processor during actual operation of the processor, and determining an actual significant temperature associated with the processor during the actual operation using the indirect thermal data for the processor during actual operation of the processor and the relationship.

  13. Calibration of Robot Reference Frames for Enhanced Robot Positioning Accuracy

    OpenAIRE

    Cheng, Frank Shaopeng

    2008-01-01

    This chapter discussed the importance and methods of conducting robot workcell calibration for enhancing the accuracy of the robot TCP positions in industrial robot applications. It shows that the robot frame transformations define the robot geometric parameters such as joint position variables, link dimensions, and joint offsets in an industrial robot system. The D-H representation allows the robot designer to model the robot motion geometry with the four standard D-H parameters. The robot k...

  14. Model Calibration in Watershed Hydrology

    Science.gov (United States)

    Yilmaz, Koray K.; Vrugt, Jasper A.; Gupta, Hoshin V.; Sorooshian, Soroosh

    2009-01-01

    Hydrologic models use relatively simple mathematical equations to conceptualize and aggregate the complex, spatially distributed, and highly interrelated water, energy, and vegetation processes in a watershed. A consequence of process aggregation is that the model parameters often do not represent directly measurable entities and must, therefore, be estimated using measurements of the system inputs and outputs. During this process, known as model calibration, the parameters are adjusted so that the behavior of the model approximates, as closely and consistently as possible, the observed response of the hydrologic system over some historical period of time. This Chapter reviews the current state-of-the-art of model calibration in watershed hydrology with special emphasis on our own contributions in the last few decades. We discuss the historical background that has led to current perspectives, and review different approaches for manual and automatic single- and multi-objective parameter estimation. In particular, we highlight the recent developments in the calibration of distributed hydrologic models using parameter dimensionality reduction sampling, parameter regularization and parallel computing.

  15. Unassisted 3D camera calibration

    Science.gov (United States)

    Atanassov, Kalin; Ramachandra, Vikas; Nash, James; Goma, Sergio R.

    2012-03-01

    With the rapid growth of 3D technology, 3D image capture has become a critical part of the 3D feature set on mobile phones. 3D image quality is affected by the scene geometry as well as on-the-device processing. An automatic 3D system usually assumes known camera poses accomplished by factory calibration using a special chart. In real life settings, pose parameters estimated by factory calibration can be negatively impacted by movements of the lens barrel due to shaking, focusing, or camera drop. If any of these factors displaces the optical axes of either or both cameras, vertical disparity might exceed the maximum tolerable margin and the 3D user may experience eye strain or headaches. To make 3D capture more practical, one needs to consider unassisted (on arbitrary scenes) calibration. In this paper, we propose an algorithm that relies on detection and matching of keypoints between left and right images. Frames containing erroneous matches, along with frames with insufficiently rich keypoint constellations, are detected and discarded. Roll, pitch yaw , and scale differences between left and right frames are then estimated. The algorithm performance is evaluated in terms of the remaining vertical disparity as compared to the maximum tolerable vertical disparity.

  16. Variability among polysulphone calibration curves

    Energy Technology Data Exchange (ETDEWEB)

    Casale, G R [University of Rome ' La Sapienza' , Physics Department, P.le A. Moro 2, I-00185, Rome (Italy); Borra, M [ISPESL - Istituto Superiore per la Prevenzione E la Sicurezza del Lavoro, Occupational Hygiene Department, Via Fontana Candida 1, I-0040 Monteporzio Catone (RM) (Italy); Colosimo, A [University of Rome ' La Sapienza' , Department of Human Physiology and Pharmacology, P.le A. Moro 2, I-00185, Rome (Italy); Colucci, M [ISPESL - Istituto Superiore per la Prevenzione E la Sicurezza del Lavoro, Occupational Hygiene Department, Via Fontana Candida 1, I-0040 Monteporzio Catone (RM) (Italy); Militello, A [ISPESL - Istituto Superiore per la Prevenzione E la Sicurezza del Lavoro, Occupational Hygiene Department, Via Fontana Candida 1, I-0040 Monteporzio Catone (RM) (Italy); Siani, A M [University of Rome ' La Sapienza' , Physics Department, P.le A. Moro 2, I-00185, Rome (Italy); Sisto, R [ISPESL - Istituto Superiore per la Prevenzione E la Sicurezza del Lavoro, Occupational Hygiene Department, Via Fontana Candida 1, I-0040 Monteporzio Catone (RM) (Italy)

    2006-09-07

    Within an epidemiological study regarding the correlation between skin pathologies and personal ultraviolet (UV) exposure due to solar radiation, 14 field campaigns using polysulphone (PS) dosemeters were carried out at three different Italian sites (urban, semi-rural and rural) in every season of the year. A polysulphone calibration curve for each field experiment was obtained by measuring the ambient UV dose under almost clear sky conditions and the corresponding change in the PS film absorbance, prior and post exposure. Ambient UV doses were measured by well-calibrated broad-band radiometers and by electronic dosemeters. The dose-response relation was represented by the typical best fit to a third-degree polynomial and it was parameterized by a coefficient multiplying a cubic polynomial function. It was observed that the fit curves differed from each other in the coefficient only. It was assessed that the multiplying coefficient was affected by the solar UV spectrum at the Earth's surface whilst the polynomial factor depended on the photoinduced reaction of the polysulphone film. The mismatch between the polysulphone spectral curve and the CIE erythemal action spectrum was responsible for the variability among polysulphone calibration curves. The variability of the coefficient was related to the total ozone amount and the solar zenith angle. A mathematical explanation of such a parameterization was also discussed.

  17. Net analyte signal calculation for multivariate calibration

    NARCIS (Netherlands)

    Ferre, J.; Faber, N.M.

    2003-01-01

    A unifying framework for calibration and prediction in multivariate calibration is shown based on the concept of the net analyte signal (NAS). From this perspective, the calibration step can be regarded as the calculation of a net sensitivity vector, whose length is the amount of net signal when the

  18. Optimal, Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard

    2002-01-01

    Reliability based code calibration is considered in this paper. It is described how the results of FORM based reliability analysis may be related to the partial safety factors and characteristic values. The code calibration problem is presented in a decision theoretical form and it is discussed how...... of reliability based code calibration of LRFD based design codes....

  19. Code Calibration as a Decision Problem

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard; Kroon, I. B.; Faber, M. H.

    1993-01-01

    Calibration of partial coefficients for a class of structures where no code exists is considered. The partial coefficients are determined such that the difference between the reliability for the different structures in the class considered and a target reliability level is minimized. Code...... calibration on a decision theoretical basis is discussed. Results from code calibration for rubble mound breakwater designs are shown....

  20. Systems and methods of eye tracking calibration

    DEFF Research Database (Denmark)

    2014-01-01

    Methods and systems to facilitate eye tracking control calibration are provided. One or more objects are displayed on a display of a device, where the one or more objects are associated with a function unrelated to a calculation of one or more calibration parameters. The one or more calibration...

  1. Perfecting the Photometric Calibration of the ACS CCD Cameras

    Science.gov (United States)

    Bohlin, Ralph C.

    2016-09-01

    Newly acquired data and improved data reduction algorithms mandate a fresh look at the absolute flux calibration of the charge-coupled device cameras on the Hubble Space Telescope (HST) Advanced Camera for Surveys (ACS). The goals are to achieve a 1% accuracy and to make this calibration more accessible to the HST guest investigator. Absolute fluxes from the CALSPEC1 database for three primary hot 30,000-60,000K WDs define the sensitivity calibrations for the Wide Field Channel (WFC) and High Resolution Channel (HRC) filters. The external uncertainty for the absolute flux is ˜1%, while the internal consistency of the sensitivities in the broadband ACS filters is ˜0.3% among the three primary WD flux standards. For stars as cool as K type, the agreement with the CALSPEC standards is within 1% at the WFC1-1K subarray position, which achieves the 1% precision goal for the first time. After making a small adjustment to the filter bandpass for F814W, the 1% precision goal is achieved over the full F814W WFC field of view for stars of K type and hotter. New encircled energies and absolute sensitivities replace the seminal results of Sirianni et al. that were published in 2005. After implementing the throughput updates, synthetic predictions of the WFC and HRC count rates for the average of the three primary WD standard stars agree with the observations to 0.1%.

  2. A recommendation for revised dose calibrator measurement procedures for 89Zr and 124I.

    Directory of Open Access Journals (Sweden)

    Bradley J Beattie

    Full Text Available Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed 18F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R using a published calibration setting number of 465. PET images of (124I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89Zr and (124I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an 18F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89Zr and (124I. These and other positron emitting radionuclides can be accurately calibrated relative to 18F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances.

  3. Design of an ultra-portable field transfer radiometer supporting automated vicarious calibration

    Science.gov (United States)

    Anderson, Nikolaus; Thome, Kurtis; Czapla-Myers, Jeffrey; Biggar, Stuart

    2015-09-01

    The University of Arizona Remote Sensing Group (RSG) began outfitting the radiometric calibration test site (RadCaTS) at Railroad Valley Nevada in 2004 for automated vicarious calibration of Earth-observing sensors. RadCaTS was upgraded to use RSG custom 8-band ground viewing radiometers (GVRs) beginning in 2011 and currently four GVRs are deployed providing an average reflectance for the test site. This measurement of ground reflectance is the most critical component of vicarious calibration using the reflectance-based method. In order to ensure the quality of these measurements, RSG has been exploring more efficient and accurate methods of on-site calibration evaluation. This work describes the design of, and initial results from, a small portable transfer radiometer for the purpose of GVR calibration validation on site. Prior to deployment, RSG uses high accuracy laboratory calibration methods in order to provide radiance calibrations with low uncertainties for each GVR. After deployment, a solar radiation based calibration has typically been used. The method is highly dependent on a clear, stable atmosphere, requires at least two people to perform, is time consuming in post processing, and is dependent on several large pieces of equipment. In order to provide more regular and more accurate calibration monitoring, the small portable transfer radiometer is designed for quick, one-person operation and on-site field calibration comparison results. The radiometer is also suited for laboratory calibration use and thus could be used as a transfer radiometer calibration standard for ground viewing radiometers of a RadCalNet site.

  4. DIGITAL CAMERA CALIBRATION USING IMAGES TAKEN FROM AN UNMANNED AERIAL VEHICLE

    Directory of Open Access Journals (Sweden)

    M. Pérez

    2012-09-01

    Full Text Available For calibrating the camera, an accurate determination of the interior orientation parameters is needed. For more accurate results, the calibration images should be taken under conditions that are similar to the field samples. The aim of this work is the establishment of an efficient and accurate digital camera calibration method to be used in particular working conditions, as it can be found with our UAV (Unmanned Aerial Vehicle photogrammetric projects. The UAV used in this work was md4-200 modelled by Microdrones. The microdrone is also equipped with a standard digital non- metric camera, the Pentax Optio A40 camera. To find out the interior orientation parameters of the digital camera, two calibration methods were done. A lab calibration based on a flat pattern and a field calibration were fulfilled. To carry out the calibration, Photomodeler Scanner software was used in both cases. The lab calibration process was completely automatic using a calibration grid. The focal length was fixed at widest angle and the network included a total of twelve images with± 90º roll angles. In order to develop the field calibration, a flight plan was programmed including a total of twelve images. In the same way as in the lab calibration, the focal length was fixed at widest angle. The field test used in the study was a flat surface located on the University of Almería campus and a set of 67 target points were placed. The calibration field area was 25 × 25 m approximately and the altitude flight over ground was 50 m. After the software processing, the camera calibration parameter values were obtained. The paper presents the process, the results and the accuracy of these calibration methods. The field calibration method reduced the final total error obtained in the previous lab calibration. Furthermore the overall RMSs obtained from both methods are similar. Therefore we will apply the field calibration results to all our photogrammetric projects in which

  5. [VMTBB-Based Spectral Radiometric Calibration of NIR Fiber Coupled Spectrometer].

    Science.gov (United States)

    Zheng, Feng; Liu, Li-ying; Liu, Xiao-xi; Li, Ye; Shi, Xiao-guang; Zhang, Guo-yu; Huan, Ke-wei

    2015-09-01

    nonlinear model is explained by small sample dataset. As the analytic results, there are obviously influences for the calibration precision among different strategies of selecting model's structural parameters. The calibration precision, which is mathematical described by standard deviation of spectral data for calibration, could be from ±0.1% to ±1%.

  6. [Calibration transfer between two FTNIR spectrophotometers using SVR].

    Science.gov (United States)

    Zhao, Long-lian; Li, Jun-hui; Zhang, Wen-juan; Wang, Jian-cai; Zhang, Lu-da

    2008-10-01

    In the present research, a set of maize powder samples was used to study the calibration transfer between two fourier transform near-infrared (FTNIR) spectrophotometers, and a method of moving window support vector regression machines (SVR) was used to correct the differences between the two instruments. Bruker Vector 22/N was referred to as "master" on which the maize protein calibration model was built. Bruker MPA was referred to as "slave" instrument. A transformation matrix was constructed based on the spectra of a sample set (for calibration transfer) measured on both instruments. After transfer, NIR spectra acquired on "slave" will appear as if they were measured on master instrument. The calibration model available for the master can then be used to predict the transformed spectra measured on the slave. The transfer parameters were computed as follows. For wavelength i, the absorbance vector obtained on the master instrument was regressed against the corresponding absorbance matrix of a spectral window obtained on the slave instrument. Method of SVR was used for regression Moving the wavelength i and corresponding window, the transfer parameter for each wavelength can be obtained. For the two FTNIR spectrophotometers, a window size of 31 wavelengths and a subset of 15 transfer samples were chosen to establish the SVR regression model between "master" and "slave". Applying the calibration model to the prediction samples after being corrected by the transfer parameters, a good transfer performance can be achieved. The correlation coefficient (r) is 0.9434, while the relative standard deviation (RSD) is 4.23%. These results suggest that the SVR method can be used to successfully transfer the calibration model for protein of maize developed on a FTNIR spectrophotometer to another.

  7. Efficiency calibration of a HPGe detector for [{sup 18}F] FDG activity measurements

    Energy Technology Data Exchange (ETDEWEB)

    Fragoso, Maria da Conceicao de Farias; Lacerda, Isabelle Viviane Batista de; Albuquerque, Antonio Morais de Sa, E-mail: mariacc05@yahoo.com.br, E-mail: isabelle.lacerda@ufpe.br, E-mail: moraisalbuquerque@hotmaiI.com [Universidade Federal de Pernambuco (DEN/UFPE), Recife, PE (Brazil). Departamento de Energia Nuclear; Oliveira, Mercia Liane de; Hazin, Clovis Abrahao; Lima, Fernando Roberto de Andrade, E-mail: mercial@cnen.gov.br, E-mail: chazin@cnen.gov.br, E-mail: falima@cnen.gov.br [Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE/CNEN-PE), Recife, PE (Brazil)

    2013-11-01

    The radionuclide {sup 18}F, in the form of flurodeoxyglucose (FDG), is the most used radiopharmaceutical for Positron Emission Tomography (PET). Due to [{sup 18}F]FDG increasing demand, it is important to ensure high quality activity measurements in the nuclear medicine practice. Therefore, standardized reference sources are necessary to calibrate of {sup 18}F measuring systems. Usually, the activity measurements are performed in re-entrant ionization chambers, also known as radionuclide calibrators. Among the existing alternatives for the standardization of radioactive sources, the method known as gamma spectrometry is widely used for short-lived radionuclides, since it is essential to minimize source preparation time. The purpose of this work was to perform the standardization of the [{sup 18}F]FDG solution by gamma spectrometry. In addition, the reference sources calibrated by this method can be used to calibrate and test the radionuclide calibrators from the Divisao de Producao de Radiofarmacos (DIPRA) of the Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE). Standard sources of {sup 152}Eu, {sup 137}Cs and {sup 68}Ge were used for the efficiency calibration of the spectrometer system. As a result, the efficiency curve as a function of energy was determined in wide energy range from 122 to 1408 keV. Reference sources obtained by this method can be used in [{sup 18}F]FDG activity measurements comparison programs for PET services localized in the Brazilian Northeast region. (author)

  8. A Path to NIST Calibrated Stars over the Dome of the Sky

    Science.gov (United States)

    Zimmer, P.; McGraw, J. T.; Zirzow, D. C.; Cramer, C.; Lykke, K.; Woodward, J. T., IV

    2016-05-01

    The UNM Measurement Astrophysics group is currently constructing and testing a mobile instrument suite that includes a multi-wavelength backscatter lidar, stellar spectroradiometer and cameras (visible and thermal infrared) that will provide real-time atmospheric transmission metadata in the column of atmosphere through which a supported telescope is observing. The design , operation and calibration of the lidar (the Facility Lidar for Astronomical Measurement of Extinction - FLAME) and spectroradiometer (the Astronomical Extinction SpectroPhotometer - AESoP) are detailed. The first task of this instrument suite will be to help create a new set of standard stars radiometrically calibrated to NIST standards. Initially this will be done for bright stars across the wavelength range 350nm to 1050nm at 1nm spectral resolution with measurement accuracy better than 1% per spectral resolution element by calibration to NIST silicon detectors. Because these standard stars will support both ground- and space-based observations, our proposed evolution of calibration begins with suitable bright optical standards and then adds measurements into the infrared. Following optical/infrared calibration of bright stars we plan to calibrate fainter stars, ultimately to V˜18, both in the optical and near infrared.

  9. Gemini Planet Imager Observational Calibrations II: Detector Performance and Calibration

    CERN Document Server

    Ingraham, Patrick; Sadakuni, Naru; Ruffio, Jean-Baptiste; Maire, Jerome; Chilcote, Jeff; Larkin, James; Marchis, Franck; Galicher, Raphael; Weiss, Jason

    2014-01-01

    The Gemini Planet Imager is a newly commissioned facility instrument designed to measure the near-infrared spectra of young extrasolar planets in the solar neighborhood and obtain imaging polarimetry of circumstellar disks. GPI's science instrument is an integral field spectrograph that utilizes a HAWAII-2RG detector with a SIDECAR ASIC readout system. This paper describes the detector characterization and calibrations performed by the GPI Data Reduction Pipeline to compensate for effects including bad/hot/cold pixels, persistence, non-linearity, vibration induced microphonics and correlated read noise.

  10. Photometric Calibrations for the SIRTF Infrared Spectrograph

    CERN Document Server

    Morris, P W; Herter, T L; Armus, L; Houck, J; Sloan, G

    2002-01-01

    The SIRTF InfraRed Spectrograph (IRS) is faced with many of the same calibration challenges that were experienced in the ISO SWS calibration program, owing to similar wavelength coverage and overlapping spectral resolutions of the two instruments. Although the IRS is up to ~300 times more sensitive and without moving parts, imposing unique calibration challenges on their own, an overlap in photometric sensitivities of the high-resolution modules with the SWS grating sections allows lessons, resources, and certain techniques from the SWS calibration programs to be exploited. We explain where these apply in an overview of the IRS photometric calibration planning.

  11. Calibration of fringe projection profilometry using an inaccurate 2D reference target

    Science.gov (United States)

    Liu, Xiaoli; Cai, Zewei; Yin, Yongkai; Jiang, Hao; He, Dong; He, Wenqi; Zhang, Zonghua; Peng, Xiang

    2017-02-01

    Recently, the camera calibration techniques based on planar reference target have been widely used in 3D fringe projection profilometry (FPP). Commonly a reference target with high-accuracy is necessary for achieving the accurate calibration. However it is costly to fabricate such a reference target, especially for those with very large or small sizes. This paper focuses on a study of calibration with an inaccurate planar target. Two major contributions are made to the FPP system calibration: (1) using the bundle adjustment strategy to optimally locate the coordinates of feature points on the planar reference target, allowing compensation for the fabrication errors, and (2) introducing a standard sphere to determine the correct scale factor, resulting in an accurate calibration results. The experimental results are also presented to demonstrate the proposed approach.

  12. Pre-flight calibration of LYRA, the solar VUV radiometer on board PROBA2

    Science.gov (United States)

    Benmoussa, A.; Dammasch, I. E.; Hochedez, J.-F.; Schühle, U.; Koller, S.; Stockman, Y.; Scholze, F.; Richter, M.; Kroth, U.; Laubis, C.; Dominique, M.; Kretzschmar, M.; Mekaoui, S.; Gissot, S.; Theissen, A.; Giordanengo, B.; Bolsee, D.; Hermans, C.; Gillotay, D.; Defise, J.-M.; Schmutz, W.

    2009-12-01

    Aims. LYRA, the Large Yield Radiometer, is a vacuum ultraviolet (VUV) solar radiometer, planned to be launched in November 2009 on the European Space Agency PROBA2, the Project for On-Board Autonomy spacecraft. Methods: The instrument was radiometrically calibrated in the radiometry laboratory of the Physikalisch-Technische Bundesanstalt (PTB) at the Berlin Electron Storage ring for SYnchroton radiation (BESSY II). The calibration was done using monochromatized synchrotron radiation at PTB's VUV and soft X-ray radiometry beamlines using reference detectors calibrated with the help of an electrical substitution radiometer as the primary detector standard. Results: A total relative uncertainty of the radiometric calibration of the LYRA instrument between 1% and 11% was achieved. LYRA will provide irradiance data of the Sun in four UV passbands and with high temporal resolution down to 10 ms. The present state of the LYRA pre-flight calibration is presented as well as the expected instrument performance.

  13. Calibration of ground-based Lidar instrument WLS7-318

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  14. Calibration of ground-based Lidar instrument WLS866-12

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  15. Calibration of ground-based lidar instrument WLS7-218

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  16. Calibration of ground-based lidar instrument WLS7-99

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  17. Calibration of ground-based Lidar instrument WLS7-283

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  18. Calibration of ground-based Lidar instrument WLS7-37

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  19. Calibration of ground-based Lidar instrument WLS7-341

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  20. Calibration of ground-based lidar instrument WLS7-219

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  1. Calibration of ground-based Lidar instrument WLS866-1

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  2. Calibration of ground-based Lidar instrument WLS7-327

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  3. Calibration of ground-based Lidar instrument WLS866-13

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  4. Calibration of ground-based Lidar instrument WLS7-218

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  5. Calibration of ground-based lidar instrument WLS7-213

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  6. Calibration of ground-based Lidar instrument WLS7-73

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  7. Calibration of ground-based Lidar instrument WLS7-222

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  8. Calibration of ground-based Lidar instrument WLS866-8

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  9. Calibration of ground-based Lidar instrument WLS7-226

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  10. Calibration of ground-based Lidar instrument WLS7-34

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  11. Calibration of ground-based Lidar instrument WLS7-369

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  12. Calibration of ground-based Lidar instrument WLS7-150

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  13. Calibration of ground-based Lidar instrument WLS866-5

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  14. Calibration of ground-based lidar instrument WLS7-151

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  15. Calibration of ground-based Lidar instrument WLS7-264

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  16. Calibration of ground-based lidar instrument WLS7-221

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  17. Calibration of ground-based Lidar instrument WLS7-91

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  18. Calibration of ground-based Lidar instrument WLS7-66

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  19. Calibration of ground-based Lidar instrument WLS7-284

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  20. A likelihood method to cross-calibrate air-shower detectors

    CERN Document Server

    Dembinski, H P; Mariş, I C; Roth, M; Veberič, D

    2016-01-01

    We present a detailed statistical treatment of the energy calibration of hybrid air-shower detectors, which combine a surface detector array and a fluorescence detector, to obtain an unbiased estimate of the calibration curve. The special features of calibration data from air showers prevent unbiased results, if a standard least-squares fit is applied to the problem. We develop a general maximum-likelihood approach, based on the detailed statistical model, to solve the problem. Our approach was developed for the Pierre Auger Observatory, but the applied principles are general and can be transferred to other air-shower experiments, even to the cross-calibration of other observables. Since our general likelihood function is expensive to compute, we derive two approximations with significantly smaller computational cost. In the recent years both have been used to calibrate data of the Pierre Auger Observatory. We demonstrate that these approximations introduce negligible bias when they are applied to simulated t...

  1. Calibration of ground-based Lidar instrument WLS866-4

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  2. Calibration of ground-based Lidar instrument WLS7-343

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  3. Calibration of ground-based Lidar instrument WLS7-377

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  4. Calibration of ground-based Lidar instrument WLS7-219

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  5. Calibration of ground-based lidar instrument WLS7-202

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  6. Calibration of ground-based Lidar instrument WLS866-3

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  7. Calibration of ground-based Lidar instrument WLS866-9

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  8. Calibration of ground-based lidar instrument WLS7-203

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  9. Calibration of ground-based lidar instrument WLS7-222

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  10. Calibration of ground-based Lidar instrument WLS7-9

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  11. Calibration of ground-based lidar instrument WLS7-204

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Courtney, Michael

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertaint......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbine at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  12. Calibration of ground-based Lidar instrument WLS7-269

    DEFF Research Database (Denmark)

    Gómez Arranz, Paula; Wagner, Rozenn

    This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement uncertain......This report presents the result of the lidar calibration performed for the given WLS7 Windcube at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  13. Vacuum Technology and Standardization-An Update

    Science.gov (United States)

    Akram, H. M.; Rashid, H.

    2011-06-01

    Vacuum technology has been vital for the progress in almost every field of modern industrial & scientific research and technological developments. Research in this field is therefore important for the rapid progress in other sophisticated technologies. The modern society require precise know-how of vacuum metrology for its complex and sophisticated manufacturing processes and research activities. Accuracy in vacuum measurements is therefore an essential need for every application. The required accuracy is achieved with the help of well-calibrated vacuum gauges and this is possible only, if there exist proper vacuum standards of required range and accuracy. In this paper, a brief review of recently developed different vacuum standards, namely Standard Mercury Manometer, Standard Volume Expansion System and Standard Orifice Flow System will be presented, employed for the calibration of low, medium and high vacuum gauges respectively. Our recently developed standards are simple in design, least in vibration & degassing rate with desired accuracy, ease of operation and cost effective.

  14. Muon Energy Calibration of the MINOS Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Miyagawa, Paul S.

    2004-09-01

    MINOS is a long-baseline neutrino oscillation experiment designed to search for conclusive evidence of neutrino oscillations and to measure the oscillation parameters precisely. MINOS comprises two iron tracking calorimeters located at Fermilab and Soudan. The Calibration Detector at CERN is a third MINOS detector used as part of the detector response calibration programme. A correct energy calibration between these detectors is crucial for the accurate measurement of oscillation parameters. This thesis presents a calibration developed to produce a uniform response within a detector using cosmic muons. Reconstruction of tracks in cosmic ray data is discussed. This data is utilized to calculate calibration constants for each readout channel of the Calibration Detector. These constants have an average statistical error of 1.8%. The consistency of the constants is demonstrated both within a single run and between runs separated by a few days. Results are presented from applying the calibration to test beam particles measured by the Calibration Detector. The responses are calibrated to within 1.8% systematic error. The potential impact of the calibration on the measurement of oscillation parameters by MINOS is also investigated. Applying the calibration reduces the errors in the measured parameters by {approx} 10%, which is equivalent to increasing the amount of data by 20%.

  15. Muon Energy Calibration of the MINOS Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Miyagawa, Paul S. [Somerville College, Oxford (United Kingdom)

    2004-01-01

    MINOS is a long-baseline neutrino oscillation experiment designed to search for conclusive evidence of neutrino oscillations and to measure the oscillation parameters precisely. MINOS comprises two iron tracking calorimeters located at Fermilab and Soudan. The Calibration Detector at CERN is a third MINOS detector used as part of the detector response calibration programme. A correct energy calibration between these detectors is crucial for the accurate measurement of oscillation parameters. This thesis presents a calibration developed to produce a uniform response within a detector using cosmic muons. Reconstruction of tracks in cosmic ray data is discussed. This data is utilized to calculate calibration constants for each readout channel of the Calibration Detector. These constants have an average statistical error of 1.8%. The consistency of the constants is demonstrated both within a single run and between runs separated by a few days. Results are presented from applying the calibration to test beam particles measured by the Calibration Detector. The responses are calibrated to within 1.8% systematic error. The potential impact of the calibration on the measurement of oscillation parameters by MINOS is also investigated. Applying the calibration reduces the errors in the measured parameters by ~ 10%, which is equivalent to increasing the amount of data by 20%.

  16. "Calibration-on-the-spot": How to calibrate an EMCCD camera from its images.

    Science.gov (United States)

    Mortensen, Kim I; Flyvbjerg, Henrik

    2016-07-06

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information.

  17. Calibration-on-the-spot”: How to calibrate an EMCCD camera from its images

    DEFF Research Database (Denmark)

    Mortensen, Kim; Flyvbjerg, Henrik

    2016-01-01

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction......-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information....

  18. Verification of L-band SAR calibration

    Science.gov (United States)

    Larson, R. W.; Jackson, P. L.; Kasischke, E.

    1985-01-01

    Absolute calibration of a digital L-band SAR system to an accuracy of better than 3 dB has been verified. This was accomplished with a calibration signal generator that produces the phase history of a point target. This signal relates calibration values to various SAR data sets. Values of radar cross-section (RCS) of reference reflectors were obtained using a derived calibration relationship for the L-band channel on the ERIM/CCRS X-C-L SAR system. Calibrated RCS values were compared to known RCS values of each reference reflector for verification and to obtain an error estimate. The calibration was based on the radar response to 21 calibrated reference reflectors.

  19. Calibration of the Cherenkov Telescope Array

    CERN Document Server

    Gaug, Markus; Berge, David; Reyes, Raquel de los; Doro, Michele; Foerster, Andreas; Maccarone, Maria Concetta; Parsons, Dan; van Eldik, Christopher

    2015-01-01

    The construction of the Cherenkov Telescope Array is expected to start soon. We will present the baseline methods and their extensions currently foreseen to calibrate the observatory. These are bound to achieve the strong requirements on allowed systematic uncertainties for the reconstructed gamma-ray energy and flux scales, as well as on the pointing resolution, and on the overall duty cycle of the observatory. Onsite calibration activities are designed to include a robust and efficient calibration of the telescope cameras, and various methods and instruments to achieve calibration of the overall optical throughput of each telescope, leading to both inter-telescope calibration and an absolute calibration of the entire observatory. One important aspect of the onsite calibration is a correct understanding of the atmosphere above the telescopes, which constitutes the calorimeter of this detection technique. It is planned to be constantly monitored with state-of-the-art instruments to obtain a full molecular and...

  20. Radio Interferometric Calibration Using The SAGE Algorithm

    CERN Document Server

    Kazemi, S; Zaroubi, S; de Bruyn, A G; Koopmans, L V E; Noordam, J

    2010-01-01

    The aim of the new generation of radio synthesis arrays such as LOFAR and SKA is to achieve much higher sensitivity, resolution and frequency coverage than what is available now. To accomplish this goal, the accuracy of the calibration techniques used is of considerable importance. Moreover, since these telescopes produce huge amounts of data, speed of convergence of calibration is a major bottleneck. The errors in calibration are due to system noise (sky and instrumental) as well as the estimation errors introduced by the calibration technique itself, which we call "solver noise". We define solver noise as the "distance" between the optimal solution, the true value of the unknowns corrupted by the system noise, and the solution obtained by calibration. We present the Space Alternating Generalized Expectation Maximization (SAGE) calibration technique, which is a modification of the Expectation Maximization algorithm, and compare its performance with the traditional Least Squares calibration based on the level...

  1. Accounting standards

    NARCIS (Netherlands)

    Stellinga, B.; Mügge, D.

    2014-01-01

    The European and global regulation of accounting standards have witnessed remarkable changes over the past twenty years. In the early 1990s, EU accounting practices were fragmented along national lines and US accounting standards were the de facto global standards. Since 2005, all EU listed companie

  2. Development of plating thickness standards. Milestone report

    Energy Technology Data Exchange (ETDEWEB)

    Russell, R.J.

    1978-01-01

    Standards which are unavailable from the National Bureau of Standards were developed to support the nondestructive measurement of plating thickness. Their fabrication, measurement, certification, and calibration-recall schedule are discussed. Reference standards that have been put into service include aluminum/Kapton, silver/copper, tin/steel, gold/silver, cadmium/Kovar, silver/iron, rhodium/copper, and gold/ceramic. 6 figures, 3 tables.

  3. Precise astronomical flux calibration and its impact on studying the nature of the dark energy

    Science.gov (United States)

    Stubbs, Christopher W.; Brown, Yorke J.

    2015-12-01

    Measurements of the luminosity of Type Ia supernovae versus redshift provided the original evidence for the accelerating expansion of the Universe and the existence of dark energy. Despite substantial improvements in survey methodology, systematic uncertainty in flux calibration dominates the error budget for this technique, exceeding both statistics and other systematic uncertainties. Consequently, any further collection of Type Ia supernova data will fail to refine the constraints on the nature of dark energy unless we also improve the state of the art in astronomical flux calibration to the order of 1%. We describe how these systematic errors arise from calibration of instrumental sensitivity, atmospheric transmission and Galactic extinction, and discuss ongoing efforts to meet the 1% precision challenge using white dwarf stars as celestial standards, exquisitely calibrated detectors as fundamental metrologic standards, and real-time atmospheric monitoring.

  4. Precise Astronomical Flux Calibration and its Impact on Studying the Nature of Dark Energy

    CERN Document Server

    Stubbs, Christopher W

    2016-01-01

    Measurements of the luminosity of type Ia supernovae vs. redshift provided the original evidence for the accelerating expansion of the Universe and the existence of dark energy. Despite substantial improvements in survey methodology, systematic uncertainty in flux calibration dominates the error budget for this technique, exceeding both statistics and other systematic uncertainties. Consequently, any further collection of type Ia supernova data will fail to refine the constraints on the nature of dark energy unless we also improve the state of the art in astronomical flux calibration to the order of 1%. We describe how these systematic errors arise from calibration of instrumental sensitivity, atmospheric transmission, and Galactic extinction, and discuss ongoing efforts to meet the 1% precision challenge using white dwarf stars as celestial standards, exquisitely calibrated detectors as fundamental metrologic standards, and real-time atmospheric monitoring.

  5. Calibration of TOB+ Thermometer's Cards

    CERN Document Server

    Banitt, Daniel

    2014-01-01

    Motivation - Under the new upgrade of the CMS detector the working temperature of the trackers had been reduced to -27 Celsius degrees. Though the thermal sensors themselves (Murata and Fenwal thermistors) are effective at these temperatures, the max1542 PLC (programmable logic controller) cards, interpreting the resistance of the thermal sensors into DC counts usable by the DCS (detector control system), are not designed for these temperatures in which the counts exceed their saturation and therefor had to be replaced. In my project I was in charge of handling the emplacement and calibration of the new PLC cards to the TOB (tracker outer barrel) control system.

  6. LHCb: Kali: the Framework for Fine Calibration of LHCb Calorimeter

    CERN Multimedia

    Belyaev, I; Puig, A; Savrina, D

    2010-01-01

    The fine calibration of the electromagnetic calorimeter of LHCb with real data requires the processing of O(10$^8$) events. They must be selected, reconstructed and analyzed in an iterative procedure to obtain optimal results.The dedicated calibration framework, Kali, within the LHCb software framework has been developed to fulfill these tasks. It consists of the two major components: a C++ module for data selection and a Python module for the extraction of calibration coefficients. The data selection component is a Gaudi-based application that allows to achieve a drastical reduction of input data sets by only storing the physically relevant data of the selected events. The obtained reduction is around 10$^3$ . The design of the module integrates well with the standard Grid tools for processing of large input data sets. The tool produces different output data formats: ROOT ntuples and femto-DST. While the ntuple allows various physics-driven algorithms to be used for calibration, namely iterative and global c...

  7. CMR Shuffler System: Passive Mode Calibration and Certification Report

    Energy Technology Data Exchange (ETDEWEB)

    Frame, Katherine C. [Los Alamos National Laboratory; Gomez, Cipriano D. [Retired CMR-OPS: OPERATIONS; Salazar, William R. [Los Alamos National Laboratory; Mayo, Douglas R. [Los Alamos National Laboratory; Vigil, Georgiana M. [Los Alamos National Laboratory; Crooks, William J. [Los Alamos National Laboratory; Stange, Sy [Los Alamos National Laboratory

    2012-07-20

    Los Alamos National Laboratory has a number of spherical confinement vessels (CVs) remaining from tests involving nuclear materials. These vessels have an inner diameter of 6 feet with 1 to 2 inch thick steel walls. The goal of the Confinement Vessel Disposition (CVD) project is to remove debris and reduce contamination inside the vessels. As debris is removed from the vessels, material will be placed in waste drums. Far-field gamma ray assay will be used to determine when a drum is nearing a {sup 239}Pu equivalent mass of less than 200 g. The drum will then be assayed using a waste drum shuffler operated in passive mode using a neutron coincidence counting method for accountability. This report focuses on the testing and calibration of the CMR waste drum shuffler in passive mode operation. Initial testing was performed to confirm previously accepted measurement parameters. The system was then calibrated using a set of weapons grade Pu (WGPu, {sup 239}Pu > 93%) oxide standards placed inside a 55 gallon drum. The calibration data ranges from Pu mass of 0.5 g to 188.9 g. The CMR waste drum shuffler has been tested and calibrated in passive mode in preparation for safeguards accountability measurements of waste drums containing material removed from CVs for the CVD project.

  8. Automatic Camera Calibration Using Active Displays of a Virtual Pattern.

    Science.gov (United States)

    Tan, Lei; Wang, Yaonan; Yu, Hongshan; Zhu, Jiang

    2017-03-27

    Camera calibration plays a critical role in 3D computer vision tasks. The most commonly used calibration method utilizes a planar checkerboard and can be done nearly fully automatically. However, it requires the user to move either the camera or the checkerboard during the capture step. This manual operation is time consuming and makes the calibration results unstable. In order to solve the above problems caused by manual operation, this paper presents a full-automatic camera calibration method using a virtual pattern instead of a physical one. The virtual pattern is actively transformed and displayed on a screen so that the control points of the pattern can be uniformly observed in the camera view. The proposed method estimates the camera parameters from point correspondences between 2D image points and the virtual pattern. The camera and the screen are fixed during the whole process; therefore, the proposed method does not require any manual operations. Performance of the proposed method is evaluated through experiments on both synthetic and real data. Experimental results show that the proposed method can achieve stable results and its accuracy is comparable to the standard method by Zhang.

  9. CTEPP STANDARD OPERATING PROCEDURE FOR PREPARATION OF SURROGATE RECOVERY STANDARD AND INTERNAL STANDARD SOLUTIONS FOR POLAR TARGET ANALYTES (SOP-5.26)

    Science.gov (United States)

    This SOP describes the method used for preparing surrogate recovery standard and internal standard solutions for the analysis of polar target analytes. It also describes the method for preparing calibration standard solutions for polar analytes used for gas chromatography/mass sp...

  10. Pulse-based internal calibration of polarimetric SAR

    DEFF Research Database (Denmark)

    Dall, Jørgen; Skou, Niels; Christensen, Erik Lintz

    1994-01-01

    Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops and devel......Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops...

  11. Tracked ultrasound calibration studies with a phantom made of LEGO bricks

    Science.gov (United States)

    Soehl, Marie; Walsh, Ryan; Rankin, Adam; Lasso, Andras; Fichtinger, Gabor

    2014-03-01

    In this study, spatial calibration of tracked ultrasound was compared by using a calibration phantom made of LEGO® bricks and two 3-D printed N-wire phantoms. METHODS: The accuracy and variance of calibrations were compared under a variety of operating conditions. Twenty trials were performed using an electromagnetic tracking device with a linear probe and three trials were performed using varied probes, varied tracking devices and the three aforementioned phantoms. The accuracy and variance of spatial calibrations found through the standard deviation and error of the 3-D image reprojection were used to compare the calibrations produced from the phantoms. RESULTS: This study found no significant difference between the measured variables of the calibrations. The average standard deviation of multiple 3-D image reprojections with the highest performing printed phantom and those from the phantom made of LEGO® bricks differed by 0.05 mm and the error of the reprojections differed by 0.13 mm. CONCLUSION: Given that the phantom made of LEGO® bricks is significantly less expensive, more readily available, and more easily modified than precision-machined N-wire phantoms, it prompts to be a viable calibration tool especially for quick laboratory research and proof of concept implementations of tracked ultrasound navigation.

  12. Improved Photometric Calibration of the SNLS and the SDSS Supernova Surveys

    CERN Document Server

    Betoule, M; Regnault, N; Cuillandre, J -C; Astier, P; Guy, J; Balland, C; Hage, P El; Hardin, D; Kessler, R; Guillou, L Le; Mosher, J; Pain, R; Rocci, P -F; Sako, M; Schahmaneche, K

    2012-01-01

    We present a combined photometric calibration of the SNLS and the SDSS supernova survey, which results from a joint effort of the SDSS and the SNLS collaborations. We deliver fluxes calibrated to the HST spectrophotometric star network for large sets of tertiary stars that cover the science fields of both surveys in all photometric bands. We also cross-calibrate directly the two surveys and demonstrate their consistency. For each survey the flat-fielding is revised based on the analysis of dithered star observations. The calibration transfer from the HST spectrophotometric standard stars to the multi-epoch tertiary standard star catalogs in the science fields follows three different paths: observations of primary standard stars with the SDSS PT telescope; observations of Landolt secondary standard stars with SNLS MegaCam instrument at CFHT; and direct observation of faint HST standard stars with MegaCam. In addition, the tertiary stars for the two surveys are cross-calibrated using dedicated MegaCam observati...

  13. An integrating sphere radiometer as a solution for high power calibrations in fibre optics

    Science.gov (United States)

    Carrasco-Sanz, Ana; Rodríguez-Barrios, Félix; Corredera, Pedro; Martín-López, Sonia; González-Herráez, Miguel; Hernanz, María Luisa

    2006-04-01

    This paper describes the design, characterization and calibration of a high power transfer standard for optical power measurements in optical fibres based on an integrating sphere radiometer. This radiometer, based on two detectors (Si and InGaAs), can measure powers between 100 nW and 10 W within the wavelength range of (400-1700) nm. The radiometer has been calibrated over the total spectral range of use against an electrically calibrated pyroelectric radiometer and different fibre laser diodes and ion lasers. The total uncertainty obtained is lower than ±1.5% for these wavelengths and power ranges (excluding the water absorption region).

  14. A Fundamental Parameter-Based Calibration Model for an Intrinsic Germanium X-Ray Fluorescence Spectrometer

    DEFF Research Database (Denmark)

    Christensen, Leif Højslet; Pind, Niels

    1982-01-01

    secondary target a number of relative calibration constants are calculated on the basis of knowledge of the irradiation geometry, the detector specifications, and tabulated fundamental physical parameters. The absolute calibration of the spectrometer is performed by measuring one pure element standard per......A matrix-independent fundamental parameter-based calibration model for an energy-dispersive X-ray fluorescence spectrometer has been developed. This model, which is part of a fundamental parameter approach quantification method, accounts for both the excitation and detection probability. For each...

  15. Calibration and efficiency curve of SANAEM ionization chamber for activity measurements.

    Science.gov (United States)

    Yeltepe, Emin; Kossert, Karsten; Dirican, Abdullah; Nähle, Ole; Niedergesäß, Christiane; Kemal Şahin, Namik

    2016-03-01

    A commercially available Fidelis ionization chamber was calibrated and assessed in PTB with activity standard solutions. The long-term stability and linearity of the system was checked. Energy-dependent efficiency curves for photons and beta particles were determined, using an iterative method in Excel™, to enable calibration factors to be calculated for radionuclides which were not used in the calibration. Relative deviations between experimental and calculated radionuclide efficiencies are of the order of 1% for most photon emitters and below 5% for pure beta emitters. The system will enable TAEK-SANAEM to provide traceable activity measurements.

  16. IFIN-HH ionization chamber calibration and its validation; electrometric system improvement.

    Science.gov (United States)

    Sahagia, M; Wätjen, A C; Luca, A; Ivan, C

    2010-01-01

    The paper presents the results obtained in the calibration of the CENTRONIC IG12/20A ionization chamber for 18 gamma ray emitters, and its improvement with a Keithley 6517A electrometer. The calibration figures were determined either directly in pA MBq(-1) units, or calculated from old units, by using the system capacity value. The calibration figures, determined with RML's standards, are compared with those deduced from the KCRV or the mean of the comparisons, and the values determined at PTB-Germany.

  17. IFIN-HH ionization chamber calibration and its validation; electrometric system improvement

    Energy Technology Data Exchange (ETDEWEB)

    Sahagia, M., E-mail: msahagia@nipne.r [' Horia Hulubei' National Institute of R and D for Physics and Nuclear Engineering, IFIN-HH, P.O. Box MG-6, Bucharest, RO-77125 (Romania); Waetjen, A.C.; Luca, A.; Ivan, C. [' Horia Hulubei' National Institute of R and D for Physics and Nuclear Engineering, IFIN-HH, P.O. Box MG-6, Bucharest, RO-77125 (Romania)

    2010-07-15

    The paper presents the results obtained in the calibration of the CENTRONIC IG12/20A ionization chamber for 18 gamma ray emitters, and its improvement with a Keithley 6517A electrometer. The calibration figures were determined either directly in pA MBq{sup -1} units, or calculated from old units, by using the system capacity value. The calibration figures, determined with RML's standards, are compared with those deduced from the KCRV or the mean of the comparisons, and the values determined at PTB-Germany.

  18. Calibration Response of the NSF EarthScope USArray Transportable Array

    Science.gov (United States)

    Vernon, F. L.; Harvey, D.; Eakins, J. A.; Busby, R. W.; Astiz, L.; Newman, R. L.; Reyes, J.

    2009-12-01

    The USArray Transportable Array (TA) is comprised of ~400 deployed s in a rolling layout across the 48 contiguous United States. To date, 760 s have been deployed with a Quanterra Q330 datalogger and either a Streckeisen STS-2, a Guralp CMG3T, or a Nanometrics Trillium 240 broadband sensor. As part of the operational procedures, a white noise calibration of at least 2 hours is performed at each after each site is certified and before each site is removed using the standard 40 sps data streams. Calibrations are also performed immediately before and after sensor or datalogger swaps. Results from these calibrations are remarkably consistent within each sensor type and match the expected response curves in normalized amplitude and phase quite well for the frequency range of 1 mHz to 16 Hz. Encouraged by these results, a special calibration of the TA was conducted by collecting an additional 200 sps 3 channel data stream for a week in August 2009. The TA network data return for the 200 sps calibration run was 98.7%. Using these new data, calibration response results will be presented for all sensors for the frequency range of 0.2 mHz to 80 Hz. and provide a statistical basis for comparison. To streamline white noise calibrations instigated by field operators, a web-based interface has been developed that allows authenticated users to (1) view the metadata and graphical output from the most recent calibrations carried out at a station and (2) send calibration commands to an individual station or group of stations. The interface uses asynchronous Javascript to dynamically send commands to the sensors via an Antelope Object Ring Buffer (ORB). In addition the interface logs the command history and provides some intelligent feedback to the client including disabling stations that are currently calibrating or have been queued for calibration.

  19. Crop physiology calibration in CLM

    Directory of Open Access Journals (Sweden)

    I. Bilionis

    2014-10-01

    Full Text Available Farming is using more terrestrial ground, as population increases and agriculture is increasingly used for non-nutritional purposes such as biofuel production. This agricultural expansion exerts an increasing impact on the terrestrial carbon cycle. In order to understand the impact of such processes, the Community Land Model (CLM has been augmented with a CLM-Crop extension that simulates the development of three crop types: maize, soybean, and spring wheat. The CLM-Crop model is a complex system that relies on a suite of parametric inputs that govern plant growth under a given atmospheric forcing and available resources. CLM-Crop development used measurements of gross primary productivity and net ecosystem exchange from AmeriFlux sites to choose parameter values that optimize crop productivity in the model. In this paper we calibrate these parameters for one crop type, soybean, in order to provide a faithful projection in terms of both plant development and net carbon exchange. Calibration is performed in a Bayesian framework by developing a scalable and adaptive scheme based on sequential Monte Carlo (SMC.

  20. Input calibration for negative originals

    Science.gov (United States)

    Tuijn, Chris

    1995-04-01

    One of the major challenges in the prepress environment consists of controlling the electronic color reproduction process such that a perfect match of any original can be realized. Whether this goal can be reached depends on many factors such as the dynamic range of the input device (scanner, camera), the color gamut of the output device (dye sublimation printer, ink-jet printer, offset), the color management software etc. The characterization of the color behavior of the peripheral devices is therefore very important. Photographs and positive transparents reflect the original scene pretty well; for negative originals, however, there is no obvious link to either the original scene or a particular print of the negative under consideration. In this paper, we establish a method to scan negatives and to convert the scanned data to a calibrated RGB space, which is known colorimetrically. This method is based on the reconstruction of the original exposure conditions (i.e., original scene) which generated the negative. Since the characteristics of negative film are quite diverse, a special calibration is required for each combination of scanner and film type.