WorldWideScience

Sample records for volume calibration methodology

  1. Calibration methodology for instruments utilized in X radiation beams, diagnostic level

    International Nuclear Information System (INIS)

    Penha, M. da; Potiens, A.; Caldas, L.V.E.

    2004-01-01

    Methodologies for the calibration of diagnostic radiology instruments were established at the Calibration Laboratory of IPEN. The methods may be used in the calibration procedures of survey meters used in radiation protection measurements (scattered radiation), instruments used in direct beams (attenuated and non attenuated beams) and quality control instruments. The established qualities are recommended by the international standards IEC 1267 and ISO 4037-3. Two ionization chambers were used as reference systems, one with a volume of 30 cm 3 for radiation protection measurements, and the other with a volume of 1 cm 3 for direct beam measurements. Both are traceable to the German Primary Laboratory of Physikalisch-Technische Bundesanstalt (PTB). In the case of calibration of quality control instruments, a non-invasive method using the measurement of the spectrum endpoint was established with a portable gamma and X-ray Intertechnique spectrometer system. The methods were applied to survey meters (radiation protection measurements), ionization chambers (direct beam measurements) and k Vp meters (invasive and non-invasive instruments). (Author)

  2. Calibration methodology for instruments utilized in X radiation beams, diagnostic level

    Energy Technology Data Exchange (ETDEWEB)

    Penha, M. da; Potiens, A.; Caldas, L.V.E. [Instituto de Pesquisas Energeticas e Nucleares, Comissao Nacional de Energia Nuclear, Sao Paulo (Brazil)]. E-mail: mppalbu@ipen.br

    2004-07-01

    Methodologies for the calibration of diagnostic radiology instruments were established at the Calibration Laboratory of IPEN. The methods may be used in the calibration procedures of survey meters used in radiation protection measurements (scattered radiation), instruments used in direct beams (attenuated and non attenuated beams) and quality control instruments. The established qualities are recommended by the international standards IEC 1267 and ISO 4037-3. Two ionization chambers were used as reference systems, one with a volume of 30 cm{sup 3} for radiation protection measurements, and the other with a volume of 1 cm{sup 3} for direct beam measurements. Both are traceable to the German Primary Laboratory of Physikalisch-Technische Bundesanstalt (PTB). In the case of calibration of quality control instruments, a non-invasive method using the measurement of the spectrum endpoint was established with a portable gamma and X-ray Intertechnique spectrometer system. The methods were applied to survey meters (radiation protection measurements), ionization chambers (direct beam measurements) and k Vp meters (invasive and non-invasive instruments). (Author)

  3. The KamLAND full-volume calibration system

    Energy Technology Data Exchange (ETDEWEB)

    Berger, B E [Department of Physics, Colorado State University, Fort Collins, Colorado 80523 (United States); Busenitz, J; Classen, T; Keefer, G; McKee, D; Piepke, A [Department of Physics and Astronomy, University of Alabama, Tuscaloosa, Alabama 35487 (United States); Decowski, M P; Elor, G; Frank, A; Freedman, S J; Fujikawa, B K; Galloway, M; Gray, F; Hsu, L; Ichimura, K; Kadel, R; Lendvai, C; O' Donnell, T [Physics Department, University of California, Berkeley and Lawrence Berkeley National Laboratory, Berkeley, California 94720 (United States); Dwyer, D A [W. K. Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, California 91125 (United States); Heeger, K M [Department of Physics, University of Wisconsin, Madison, Wisconsin 53706 (United States)] (and others)

    2009-04-15

    We have successfully built and operated a source deployment system for the KamLAND detector. This system was used to position radioactive sources throughout the delicate 1-kton liquid scintillator volume, while meeting stringent material cleanliness, material compatibility, and safety requirements. The calibration data obtained with this device were used to fully characterize detector position and energy reconstruction biases. As a result, the uncertainty in the size of the detector fiducial volume was reduced by a factor of two. Prior to calibration with this system, the fiducial volume was the largest source of systematic uncertainty in measuring the number of antineutrinos detected by KamLAND. This paper describes the design, operation and performance of this unique calibration system.

  4. The KamLAND Full-Volume Calibration System

    Energy Technology Data Exchange (ETDEWEB)

    KamLAND Collaboration; Berger, B. E.; Busenitz, J.; Classen, T.; Decowski, M. P.; Dwyer, D. A.; Elor, G.; Frank, A.; Freedman, S. J.; Fujikawa, B. K.; Galloway, M.; Gray, F.; Heeger, K. M.; Hsu, L.; Ichimura, K.; Kadel, R.; Keefer, G.; Lendvai, C.; McKee, D.; O' Donnell, T.; Piepke, A.; Steiner, H. M.; Syversrud, D.; Wallig, J.; Winslow, L. A.; Ebihara, T.; Enomoto, S.; Furuno, K.; Gando, Y.; Ikeda, H.; Inoue, K.; Kibe, Y.; Kishimoto, Y.; Koga, M.; Minekawa, Y.; Mitsui, T.; Nakajima, K.; Nakajima, K.; Nakamura, K.; Owada, K.; Shimizu, I.; Shimizu, Y.; Shirai, J.; Suekane, F.; Suzuki, A.; Tamae, K.; Yoshida, S.; Kozlov, A.; Murayama, H.; Grant, C.; Leonard, D. S.; Luk, K.-B.; Jillings, C.; Mauger, C.; McKeown, R. D.; Zhang, C.; Lane, C. E.; Maricic, J.; Miletic, T.; Batygov, M.; Learned, J. G.; Matsuno, S.; Pakvasa, S.; Foster, J.; Horton-Smith, G. A.; Tang, A.; Dazeley, S.; Downum, K. E.; Gratta, G.; Tolich, K.; Bugg, W.; Efremenko, Y.; Kamyshkov, Y.; Perevozchikov, O.; Karwowski, H. J.; Markoff, D. M.; Tornow, W.; Piquemal, F.; Ricol, J.-S.

    2009-03-05

    We have successfully built and operated a source deployment system for the KamLAND detector. This system was used to position radioactive sources throughout the delicate 1-kton liquid scintillator volume, while meeting stringent material cleanliness, material compatibility, and safety requirements. The calibration data obtained with this device were used to fully characterize detector position and energy reconstruction biases. As a result, the uncertainty in the size of the detector fiducial volume was reduced by a factor of two. Prior to calibration with this system, the fiducial volume was the largest source of systematic uncertainty in measuring the number of antineutrinos detected by KamLAND. This paper describes the design, operation and performance of this unique calibration system.

  5. Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry

    International Nuclear Information System (INIS)

    Brady, S. L.; Kaufman, R. A.

    2012-01-01

    Purpose: The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by ∼25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. Methods: The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Results: Calibration precision was measured to be better than 5%–7%, 3%–5%, and 2%–4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.7 ± 1.1 mV cGy −1 versus the CT scatter phantom 29.2 ± 1.0 mV cGy −1 and FIA with x-ray 29.9 ± 1.1 mV cGy −1 methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of ∼3000 mV. Conclusions: The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration coefficients for the

  6. Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry.

    Science.gov (United States)

    Brady, S L; Kaufman, R A

    2012-06-01

    The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by ~25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Calibration precision was measured to be better than 5%-7%, 3%-5%, and 2%-4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.7 ± 1.1 mV cGy(-1) versus the CT scatter phantom 29.2 ± 1.0 mV cGy(-1) and FIA with x-ray 29.9 ± 1.1 mV cGy(-1) methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of ~3000 mV. The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration coefficients for the eventual use for phantom dosimetry, a measurement error ~12

  7. Computerized 50 liter volume calibration system

    International Nuclear Information System (INIS)

    Proffitt, T.H.

    1990-01-01

    A system has been designed for the Savannah River Site that will be used to calibrate product shipping containers. For accountability purposes, it is necessary that these containers be calibrated to a very high precision. The Computerized 50 Liter Volume Calibration System (CVCS), which is based on the Ideal Gas Law (IGL), will use reference volumes with precision of no less ±0.03%, and helium to calibrate the containers to have a total error of no greater than ±0.10%. A statistical interpretation of the system has given a theoretical total calculated error of ±0.08%. Tests with the system will be performed once fabrication is complete to experimentally verify the calculated error. Since the total error was calculated using the worst case scenario, the actual error should be significantly less than the calculated value. The computer controlled, totally automated system is traceable to the National Institute of Standards and Technology. The design, calibration procedure, and statistical interpretation of the system will be discussed. 1 ref

  8. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 2: Data standardization for tank calibration

    International Nuclear Information System (INIS)

    2007-01-01

    Measurements of the volume and height of liquid in a process accountancy tank are often made in order to estimate or verify the tank's calibration or volume measurement equation. The calibration equation relates the response of the tank's measurement system to some independent measure of tank volume. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation (the inverse of the calibration equation), which relates tank volume to measurement system response. In this part of ISO 18213, it is assumed that the primary measurement-system response variable is liquid height and that the primary measure of liquid content is volume. This part of ISO 18213 presents procedures for standardizing a set of calibration data to a fixed set of reference conditions so as to minimize the effect of variations in ambient conditions that occur during the measurement process. The procedures presented herein apply generally to measurements of liquid height and volume obtained for the purpose of calibrating a tank (i.e. calibrating a tank's measurement system). When used in connection with other parts of ISO 18213, these procedures apply specifically to tanks equipped with bubbler probe systems for measuring liquid content. The standardization algorithms presented herein can be profitably applied when only estimates of ambient conditions, such as temperature, are available. However, the most reliable results are obtained when relevant ambient conditions are measured for each measurement of volume and liquid height in a set of calibration data. Information is provided on scope, physical principles, data required, calibration data, dimensional changes in the tank, multiple calibration runs and results on standardized calibration data. Four annexes inform about density of water, buoyancy corrections for mass determination, determination of tank heel volume and statistical method for aligning data from several calibration runs. A bibliography is

  9. Calibration of a large volume argon-41 gas-effluent monitor

    International Nuclear Information System (INIS)

    Wilson, William E.; Lovas, Thomas A.

    1976-01-01

    In September of 1975, a large volume Argon-41 sampler was calibrated using a series connected calibration chamber of known sensitivity and a constant flow of activated Argon gas. The calibration included analysis of the effects of flow rate through the large volume sampler and yielded a calibration constant of 2.34 x 10 -8 μc/cm 3 /CPM. (author)

  10. Gravimetric gas determinations for volume calibration

    International Nuclear Information System (INIS)

    Gibbs, P.W.

    1991-01-01

    Gravimetric measurements of gases is one of the methods available for calibrating gas volumes. By inputting a known quantity of gas and measuring the resulting pressure and temperature, the system volume can be calculated using gas law principles. Historically, this method has been less accurate due to the difficulty in the mass determination. This difficulty comes from several sources. Two examples are the large tare weight of the gas container relative to the weight of gas and the external volume of the gas container relative to the standards. The application of a gravimetric gas determination to tank volume calibrations at the savannah River Site is discussed. Mass determinations on a 25,00 gram gas container were such that a 1500 gram quantity of gas was routinely determined to within ±0.2 gram at the 99% confidence level. In this paper the weighting design and the methods used to address the difficulties of the mass determination are detailed

  11. Fuel conditioning facility electrorefiner volume calibration

    International Nuclear Information System (INIS)

    Bucher, R.G.; Orechwa, Y.

    1995-01-01

    In one of the electrometallurgical process steps of the Fuel Conditioning Facility (FCF), die in-process nuclear material is dissolved in the electrorefiner tank in an upper layer of a mixture of liquid LiCl-KCl salt and a lower layer of liquid cadmium. The electrorefiner tank, as most process tanks, is not a smooth right-circular cylinder for which a single linear volume calibration curve could be fitted over the whole height of the tank. Rather, the tank contains many internal components, which cause systematic deviations from a single linear function. The nominal operating temperature of the electrorefiner is 500 degrees C although the salt and cadmium are introduced at 410 degrees C. The operating materials and temperatures preclude multiple calibration runs at operating conditions. In order to maximize the calibration information, multiple calibration runs were performed with water at room temperature. These data allow identification of calibration segments, and preliminary estimation of the calibration function and calibration uncertainties. The final calibration function is based on a combination of data from die water calibrations and the measurements made during the filling of the electrorefiner with salt and cadmium for operation

  12. Volume calibration for nuclear materials control: ANSI N15.19-1989 and beyond

    International Nuclear Information System (INIS)

    Liebetrau, A.M.

    1994-03-01

    Since the last IAEA International Safeguards Symposium, a revised standard for volume calibration methodology was issued in the United States. Because the new standard reflects the advent of high-precision volume measurement technology, it is significantly different from the earlier standard which it supersedes. The new standard outlines a unified data standardization model that applies to process tanks equipped with differential pressure measurement systems for determining liquid content. At the heart of the model is an algorithm to determine liquid height from pressure measurements that accounts for the major factors affecting the accuracy of those measurements. The standardization model also contains algorithms that adjust data from volumetric and gravimetric provers to a standard set of reference conditions. A key component of the standardization model is an algorithm to take account of temperature-induced dimensional changes in the tank. Improved methods for the statistical treatment of calibration data have also appeared since the last Safeguards Symposium. A statistical method of alignment has been introduced that employs a least-squares criterion to determine ''optimal'' alignment factors. More importantly, a statistical model has been proposed that yields plausible estimates of the variance of height and volume measurements when significant run-to-run differences are present in the calibration data. The new standardization model and statistical methods described here are being implemented in a portable, user-friendly software program for use by IAEA inspectors and statisticians. Perhaps these methods will eventually find their way into appropriate international standards

  13. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    International Nuclear Information System (INIS)

    Cornic, Philippe; Le Besnerais, Guy; Champagnat, Frédéric; Illoul, Cédric; Cheminet, Adam; Le Sant, Yves; Leclaire, Benjamin

    2016-01-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data. (paper)

  14. Syringe calibration factors and volume correction factors for the NPL secondary standard radionuclide calibrator

    CERN Document Server

    Tyler, D K

    2002-01-01

    The activity assay of a radiopharmaceutical administration to a patient is normally achieved via the use of a radionuclide calibrator. Because of the different geometries and elemental compositions between plastic syringes and glass vials, the calibration factors for syringes may well be significantly different from those for the glass containers. The magnitude of these differences depends on the energies of the emitted photons. For some radionuclides variations have been observed of 70 %, it is therefore important to recalibrate for syringes or use syringe calibration factors. Calibration factors and volume correction factors have been derived for the NPL secondary standard radionuclide calibrator, for a variety of commonly used syringes and needles, for the most commonly used medical radionuclide.

  15. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 1: Procedural overview

    International Nuclear Information System (INIS)

    2007-01-01

    Accurate determinations of volume are a fundamental component of any measurement-based system of control and accountability in a facility that processes or stores nuclear materials in liquid form. Volume determinations are typically made with the aid of a calibration or volume measurement equation that relates the response of the tank's measurement system to some independent measure of tank volume. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation (the inverse of the calibration equation), which relates tank volume to measurement system response. The steps carried out to acquire data for estimating the tank's calibration or volume measurement equation are collectively described as the process of tank calibration. This part of ISO 18213 describes procedures for tank calibration and volume determination for nuclear process tanks equipped with pressure-measurement systems for determining liquid content. Specifically, overall guidance is provided for planning a calibration exercise undertaken to obtain the data required for the measurement equation to estimate a tank's volume. The key steps in the procedure are also presented for subsequently using the estimated volume-measurement equation to determine tank liquid volumes. The procedures presented apply specifically to tanks equipped with bubbler probe systems for measuring liquid content. Moreover, these procedures produce reliable results only for clear (i.e. without suspended solids), homogeneous liquids that are at both thermal and static equilibrium. The paper elaborates on scope, physical principles involved, the calibration model, equipment required, a typical tank calibration procedure, calibration planning and pre-calibration activities, and volume determination. A bibliography is provided

  16. Calibration methodology for proportional counters applied to yield measurements of a neutron burst

    International Nuclear Information System (INIS)

    Tarifeño-Saldivia, Ariel; Pavez, Cristian; Soto, Leopoldo; Mayer, Roberto E

    2015-01-01

    This work introduces a methodology for the yield measurement of a neutron burst using neutron proportional counters. The methodology is based on the calibration of the counter in pulse mode, and the use of a statistical model to estimate the number of detected events from the accumulated charge resulting from detection of the burst of neutrons. An improvement of more than one order of magnitude in the accuracy of a paraffin wax moderated 3 He-filled tube is obtained by using this methodology with respect to previous calibration methods. (paper)

  17. PLEIADES ABSOLUTE CALIBRATION : INFLIGHT CALIBRATION SITES AND METHODOLOGY

    Directory of Open Access Journals (Sweden)

    S. Lachérade

    2012-07-01

    Full Text Available In-flight calibration of space sensors once in orbit is a decisive step to be able to fulfil the mission objectives. This article presents the methods of the in-flight absolute calibration processed during the commissioning phase. Four In-flight calibration methods are used: absolute calibration, cross-calibration with reference sensors such as PARASOL or MERIS, multi-temporal monitoring and inter-bands calibration. These algorithms are based on acquisitions over natural targets such as African deserts, Antarctic sites, La Crau (Automatic calibration station and Oceans (Calibration over molecular scattering or also new extra-terrestrial sites such as the Moon and selected stars. After an overview of the instrument and a description of the calibration sites, it is pointed out how each method is able to address one or several aspects of the calibration. We focus on how these methods complete each other in their operational use, and how they help building a coherent set of information that addresses all aspects of in-orbit calibration. Finally, we present the perspectives that the high level of agility of PLEIADES offers for the improvement of its calibration and a better characterization of the calibration sites.

  18. Calibration methodology for proportional counters applied to yield measurements of a neutron burst

    Energy Technology Data Exchange (ETDEWEB)

    Tarifeño-Saldivia, Ariel, E-mail: atarifeno@cchen.cl, E-mail: atarisal@gmail.com; Pavez, Cristian; Soto, Leopoldo [Comisión Chilena de Energía Nuclear, Casilla 188-D, Santiago (Chile); Center for Research and Applications in Plasma Physics and Pulsed Power, P4, Santiago (Chile); Departamento de Ciencias Fisicas, Facultad de Ciencias Exactas, Universidad Andres Bello, Republica 220, Santiago (Chile); Mayer, Roberto E. [Instituto Balseiro and Centro Atómico Bariloche, Comisión Nacional de Energía Atómica and Universidad Nacional de Cuyo, San Carlos de Bariloche R8402AGP (Argentina)

    2014-01-15

    This paper introduces a methodology for the yield measurement of a neutron burst using neutron proportional counters. This methodology is to be applied when single neutron events cannot be resolved in time by nuclear standard electronics, or when a continuous current cannot be measured at the output of the counter. The methodology is based on the calibration of the counter in pulse mode, and the use of a statistical model to estimate the number of detected events from the accumulated charge resulting from the detection of the burst of neutrons. The model is developed and presented in full detail. For the measurement of fast neutron yields generated from plasma focus experiments using a moderated proportional counter, the implementation of the methodology is herein discussed. An experimental verification of the accuracy of the methodology is presented. An improvement of more than one order of magnitude in the accuracy of the detection system is obtained by using this methodology with respect to previous calibration methods.

  19. Calibration methodology for proportional counters applied to yield measurements of a neutron burst

    International Nuclear Information System (INIS)

    Tarifeño-Saldivia, Ariel; Pavez, Cristian; Soto, Leopoldo; Mayer, Roberto E.

    2014-01-01

    This paper introduces a methodology for the yield measurement of a neutron burst using neutron proportional counters. This methodology is to be applied when single neutron events cannot be resolved in time by nuclear standard electronics, or when a continuous current cannot be measured at the output of the counter. The methodology is based on the calibration of the counter in pulse mode, and the use of a statistical model to estimate the number of detected events from the accumulated charge resulting from the detection of the burst of neutrons. The model is developed and presented in full detail. For the measurement of fast neutron yields generated from plasma focus experiments using a moderated proportional counter, the implementation of the methodology is herein discussed. An experimental verification of the accuracy of the methodology is presented. An improvement of more than one order of magnitude in the accuracy of the detection system is obtained by using this methodology with respect to previous calibration methods

  20. Personal dosimetry service of TECNATOM: measurement system and methodology of calibration

    International Nuclear Information System (INIS)

    Marchena, Paloma; Bravo, Borja

    2008-01-01

    Full text: The implementation of a new integrated and practical working tool called ALEDIN within the Personal Dosimetry Service (PDS) of TECNATOM, have harmonized the methodology for the counting acquisition, detector calibration and data analysis using a friendly Windows (registered mark) environment. The knowledge of this methodology, due to the fact that is the final product of a R and D project, will help the users and the Regulatory Body for a better understanding of the internal activity measurement in individuals, allowing a more precise error identification and correction, and improving the whole process of the internal dosimetry. The development and implementation of a new calibration system of the whole body counters using NaI (Tl) detectors and the utilization of a new humanoid anthropometric phantom, BOMAB type, with a uniform radioactive source distributions, allow a better energy and activity calibration for different counting geometries covering a wide range of gamma spectra from low energies, less than 100 keV to about 2000 keV for the high energies spectra. This new calibration methodology implied the development of an improved system for the determination of the isotopic activity. This new system has been integrated in a Windows (registered mark) environment, applicable for counting acquisition and data analysis in the whole body counters WBC in cross connection with the INDAC software, which allow the interpretation of the measured activity as committed effective dose following all the new ICRP recommendations and dosimetric models for internal dose and bioassay measurements. (author)

  1. A methodology to calibrate water saturation estimated from 4D seismic data

    International Nuclear Information System (INIS)

    Davolio, Alessandra; Maschio, Célio; José Schiozer, Denis

    2014-01-01

    Time-lapse seismic data can be used to estimate saturation changes within a reservoir, which is valuable information for reservoir management as it plays an important role in updating reservoir simulation models. The process of updating reservoir properties, history matching, can incorporate estimated saturation changes qualitatively or quantitatively. For quantitative approaches, reliable information from 4D seismic data is important. This work proposes a methodology to calibrate the volume of water in the estimated saturation maps, as these maps can be wrongly estimated due to problems with seismic signals (such as noise, errors associated with data processing and resolution issues). The idea is to condition the 4D seismic data to known information provided by engineering, in this case the known amount of injected and produced water in the field. The application of the proposed methodology in an inversion process (previously published) that estimates saturation from 4D seismic data is presented, followed by a discussion concerning the use of such data in a history matching process. The methodology is applied to a synthetic dataset to validate the results, the main of which are: (1) reduction of the effects of noise and errors in the estimated saturation, yielding more reliable data to be used quantitatively or qualitatively and (2) an improvement in the properties update after using this data in a history matching procedure. (paper)

  2. Left ventricular volume measurement in mice by conductance catheter: evaluation and optimization of calibration

    DEFF Research Database (Denmark)

    Nielsen, Jan Møller; Kristiansen, Steen B; Ringgaard, Steffen

    2007-01-01

    in mice (n = 52) with a Millar CC (SPR-839) and compared with MRI-derived volumes (V(MRI)). Significant correlations between V(CC) and V(MRI) [end-diastolic volume (EDV): R(2) = 0.85, P 2) = 0.88, P ... in the pulmonary artery was used to calibrate for parallel conductance and volume conversion was done by individual cylinder calibration. However, a significant underestimation was observed [EDV = -17.3 microl (-22.7 to -11.9 microl); ESV = -8.8 microl (-12.5 to -5.1 microl)]. Intravenous injection....... The dual-frequency method for estimation of parallel conductance failed to produce V(CC) that correlated with V(MRI). We conclude that selection of the calibration procedure for the CC has significant implications for the accuracy and precision of volume estimation and pressure-volume loop...

  3. Methodology for generating waste volume estimates

    International Nuclear Information System (INIS)

    Miller, J.Q.; Hale, T.; Miller, D.

    1991-09-01

    This document describes the methodology that will be used to calculate waste volume estimates for site characterization and remedial design/remedial action activities at each of the DOE Field Office, Oak Ridge (DOE-OR) facilities. This standardized methodology is designed to ensure consistency in waste estimating across the various sites and organizations that are involved in environmental restoration activities. The criteria and assumptions that are provided for generating these waste estimates will be implemented across all DOE-OR facilities and are subject to change based on comments received and actual waste volumes measured during future sampling and remediation activities. 7 figs., 8 tabs

  4. Calibration Modeling Methodology to Optimize Performance for Low Range Applications

    Science.gov (United States)

    McCollum, Raymond A.; Commo, Sean A.; Parker, Peter A.

    2010-01-01

    Calibration is a vital process in characterizing the performance of an instrument in an application environment and seeks to obtain acceptable accuracy over the entire design range. Often, project requirements specify a maximum total measurement uncertainty, expressed as a percent of full-scale. However in some applications, we seek to obtain enhanced performance at the low range, therefore expressing the accuracy as a percent of reading should be considered as a modeling strategy. For example, it is common to desire to use a force balance in multiple facilities or regimes, often well below its designed full-scale capacity. This paper presents a general statistical methodology for optimizing calibration mathematical models based on a percent of reading accuracy requirement, which has broad application in all types of transducer applications where low range performance is required. A case study illustrates the proposed methodology for the Mars Entry Atmospheric Data System that employs seven strain-gage based pressure transducers mounted on the heatshield of the Mars Science Laboratory mission.

  5. Calibration methodology for energy management system of a plug-in hybrid electric vehicle

    International Nuclear Information System (INIS)

    Duan, Benming; Wang, Qingnian; Zeng, Xiaohua; Gong, Yinsheng; Song, Dafeng; Wang, Junnian

    2017-01-01

    Highlights: • Calibration theory of EMS is proposed. • A comprehensive evaluating indicator is constructed by radar chart method. • Optimal Latin hypercube design algorithm is introduced to obtain training data. • An approximation model is established by using a RBF neural network. • Offline calibration methodology improves the actual calibration efficiency. - Abstract: This paper presents a new analytical calibration method for energy management strategy designed for a plug-in hybrid electric vehicle. This method improves the actual calibration efficiency to reach a compromise among the conflicting calibration requirements (e.g. emissions and economy). A comprehensive evaluating indicator covering emissions and economic performance is constructed by using a radar chart method. A radial basis functions (RBFs) neural network model is proposed to establish a precise model among control parameters and the comprehensive evaluation indicator. The optimal Latin hypercube design is introduced to obtain the experimental data to train the RBFs neural network model. And multi-island genetic algorithm is used to solve the optimization model. Finally, an offline calibration example is conducted. Results validate the effectiveness of the proposed calibration approach in improving vehicle performance and calibration efficiency.

  6. CONTAMINATED SOIL VOLUME ESTIMATE TRACKING METHODOLOGY

    International Nuclear Information System (INIS)

    Durham, L.A.; Johnson, R.L.; Rieman, C.; Kenna, T.; Pilon, R.

    2003-01-01

    The U.S. Army Corps of Engineers (USACE) is conducting a cleanup of radiologically contaminated properties under the Formerly Utilized Sites Remedial Action Program (FUSRAP). The largest cost element for most of the FUSRAP sites is the transportation and disposal of contaminated soil. Project managers and engineers need an estimate of the volume of contaminated soil to determine project costs and schedule. Once excavation activities begin and additional remedial action data are collected, the actual quantity of contaminated soil often deviates from the original estimate, resulting in cost and schedule impacts to the project. The project costs and schedule need to be frequently updated by tracking the actual quantities of excavated soil and contaminated soil remaining during the life of a remedial action project. A soil volume estimate tracking methodology was developed to provide a mechanism for project managers and engineers to create better project controls of costs and schedule. For the FUSRAP Linde site, an estimate of the initial volume of in situ soil above the specified cleanup guidelines was calculated on the basis of discrete soil sample data and other relevant data using indicator geostatistical techniques combined with Bayesian analysis. During the remedial action, updated volume estimates of remaining in situ soils requiring excavation were calculated on a periodic basis. In addition to taking into account the volume of soil that had been excavated, the updated volume estimates incorporated both new gamma walkover surveys and discrete sample data collected as part of the remedial action. A civil survey company provided periodic estimates of actual in situ excavated soil volumes. By using the results from the civil survey of actual in situ volumes excavated and the updated estimate of the remaining volume of contaminated soil requiring excavation, the USACE Buffalo District was able to forecast and update project costs and schedule. The soil volume

  7. Volume corrections factors in the measurement of 99mTc and 123I activities in radionuclide calibrators

    International Nuclear Information System (INIS)

    Correia, Amanda Ribeiro; Rezende, Eduarda Alexandre; Iwahara, Akira; Oliveira, Antonio Eduardo de; Oliveira, Estela Maria de; Tauhata, Luiz; Chaves, Taina Olivieri

    2012-01-01

    To determine correction factors for the variation in volume of radiopharmaceuticals in containers of different geometries, comparing the influence of such factors on the determination of 99m Tc and 123 I activity with two types of calibrators - one with ionization chamber and another with Geiger-Mueller (G-M) detector -; and to evaluate calibrators performance in the measurement of 99m Tc and 1 '2 3 I activities. Materials and Methods: Eight calibrators, 10R glass vials, 3 and 5 mL plastic syringes and 99m Tc and 123 I solutions were utilized. The correction factors were determined with basis on practical measurements of the variation in the calibrators' response according to the volume of radionuclide solution in the glass vials. The performance was evaluated according to the acceptance criterion of +- 10% accuracy required by the Brazilian standard. Results: The variation of the calibrators' response according to the variation in radionuclide volume was reasonably greater in the calibrator with G-M detector. It was also greater for 123 I than for 99m Tc. Conclusion: The results confirm that the calibrators' response depends on the radionuclide volume contained in the vials. Such dependence is more critical for the calibrators equipped with G-M detector and for 123 I as compared with 99m Tc. (author)

  8. A methodology to develop computational phantoms with adjustable posture for WBC calibration

    International Nuclear Information System (INIS)

    Fonseca, T C Ferreira; Vanhavere, F; Bogaerts, R; Hunt, John

    2014-01-01

    A Whole Body Counter (WBC) is a facility to routinely assess the internal contamination of exposed workers, especially in the case of radiation release accidents. The calibration of the counting device is usually done by using anthropomorphic physical phantoms representing the human body. Due to such a challenge of constructing representative physical phantoms a virtual calibration has been introduced. The use of computational phantoms and the Monte Carlo method to simulate radiation transport have been demonstrated to be a worthy alternative. In this study we introduce a methodology developed for the creation of realistic computational voxel phantoms with adjustable posture for WBC calibration. The methodology makes use of different software packages to enable the creation and modification of computational voxel phantoms. This allows voxel phantoms to be developed on demand for the calibration of different WBC configurations. This in turn helps to study the major source of uncertainty associated with the in vivo measurement routine which is the difference between the calibration phantoms and the real persons being counted. The use of realistic computational phantoms also helps the optimization of the counting measurement. Open source codes such as MakeHuman and Blender software packages have been used for the creation and modelling of 3D humanoid characters based on polygonal mesh surfaces. Also, a home-made software was developed whose goal is to convert the binary 3D voxel grid into a MCNPX input file. This paper summarizes the development of a library of phantoms of the human body that uses two basic phantoms called MaMP and FeMP (Male and Female Mesh Phantoms) to create a set of male and female phantoms that vary both in height and in weight. Two sets of MaMP and FeMP phantoms were developed and used for efficiency calibration of two different WBC set-ups: the Doel NPP WBC laboratory and AGM laboratory of SCK-CEN in Mol, Belgium. (paper)

  9. A methodology to develop computational phantoms with adjustable posture for WBC calibration

    Science.gov (United States)

    Ferreira Fonseca, T. C.; Bogaerts, R.; Hunt, John; Vanhavere, F.

    2014-11-01

    A Whole Body Counter (WBC) is a facility to routinely assess the internal contamination of exposed workers, especially in the case of radiation release accidents. The calibration of the counting device is usually done by using anthropomorphic physical phantoms representing the human body. Due to such a challenge of constructing representative physical phantoms a virtual calibration has been introduced. The use of computational phantoms and the Monte Carlo method to simulate radiation transport have been demonstrated to be a worthy alternative. In this study we introduce a methodology developed for the creation of realistic computational voxel phantoms with adjustable posture for WBC calibration. The methodology makes use of different software packages to enable the creation and modification of computational voxel phantoms. This allows voxel phantoms to be developed on demand for the calibration of different WBC configurations. This in turn helps to study the major source of uncertainty associated with the in vivo measurement routine which is the difference between the calibration phantoms and the real persons being counted. The use of realistic computational phantoms also helps the optimization of the counting measurement. Open source codes such as MakeHuman and Blender software packages have been used for the creation and modelling of 3D humanoid characters based on polygonal mesh surfaces. Also, a home-made software was developed whose goal is to convert the binary 3D voxel grid into a MCNPX input file. This paper summarizes the development of a library of phantoms of the human body that uses two basic phantoms called MaMP and FeMP (Male and Female Mesh Phantoms) to create a set of male and female phantoms that vary both in height and in weight. Two sets of MaMP and FeMP phantoms were developed and used for efficiency calibration of two different WBC set-ups: the Doel NPP WBC laboratory and AGM laboratory of SCK-CEN in Mol, Belgium.

  10. Calibration of a dedicated software for 3D rendering

    Energy Technology Data Exchange (ETDEWEB)

    Abrantes, Marcos E.S.; Felix, Warley F.; Veloso, Maria Auxiliadora F., E-mail: marcos.nuclear@yahoo.com.br, E-mail: warleyferreirafelix@gmail.com, E-mail: mdora@nuclear.ufmg.br [Faculdade Ciencias Medicas de Minas Gerais (FCMMG), Belo Horizonte, MG (Brazil); Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear

    2017-11-01

    With the increasing use of 3D reconstruction techniques, to assist in diagnosis, dedicated programs are being widely used. For this they must be calibrated in order to encounter the values of the real volumes of the human tissues. The purpose of this work is to indicate correction and calibration values for true volumes, read in a 3D reconstruction system dedicated, using DICOM images of Computed Tomography. This work utilized a PMMA thorax phantom associated with the DICOM image and the volume found by a program of a tomograph. The physical volume of the PMMA phantom found was 10359.0 cm³. For the volumes found according to the structures of interest, the values are 11005.5 cm³, 10249.3 cm³ and 10205.1 cm³ and the correction values are -6.2%, +1.1% e +1.5% respectively for tissues: pulmonary, bony and soft tissues. The procedure performed can be used for calibration in other 3D reconstruction programs, observing the necessary corrections and the methodology used. (author)

  11. Calibration of a dedicated software for 3D rendering

    International Nuclear Information System (INIS)

    Abrantes, Marcos E.S.; Felix, Warley F.; Veloso, Maria Auxiliadora F.; Universidade Federal de Minas Gerais

    2017-01-01

    With the increasing use of 3D reconstruction techniques, to assist in diagnosis, dedicated programs are being widely used. For this they must be calibrated in order to encounter the values of the real volumes of the human tissues. The purpose of this work is to indicate correction and calibration values for true volumes, read in a 3D reconstruction system dedicated, using DICOM images of Computed Tomography. This work utilized a PMMA thorax phantom associated with the DICOM image and the volume found by a program of a tomograph. The physical volume of the PMMA phantom found was 10359.0 cm³. For the volumes found according to the structures of interest, the values are 11005.5 cm³, 10249.3 cm³ and 10205.1 cm³ and the correction values are -6.2%, +1.1% e +1.5% respectively for tissues: pulmonary, bony and soft tissues. The procedure performed can be used for calibration in other 3D reconstruction programs, observing the necessary corrections and the methodology used. (author)

  12. New method of assigning uncertainty in volume calibration

    International Nuclear Information System (INIS)

    Lechner, J.A.; Reeve, C.P.; Spiegelman, C.H.

    1980-12-01

    This paper presents a practical statistical overview of the pressure-volume calibration curve for large nuclear materials processing tanks. It explains the appropriateness of applying splines (piecewise polynomials) to this curve, and it presents an overview of the associated statistical uncertainties. In order to implement these procedures, a practical and portable FORTRAN IV program is provided along with its users' manual. Finally, the recommended procedure is demonstrated on actual tank data collected by NBS

  13. Development of a calibration methodology and tests of kerma area product meters

    International Nuclear Information System (INIS)

    Costa, Nathalia Almeida

    2013-01-01

    The quantity kerma area product (PKA) is important to establish reference levels in diagnostic radiology exams. This quantity can be obtained using a PKA meter. The use of such meters is essential to evaluate the radiation dose in radiological procedures and is a good indicator to make sure that the dose limit to the patient's skin doesn't exceed. Sometimes, these meters come fixed to X radiation equipment, which makes its calibration difficult. In this work, it was developed a methodology for calibration of PKA meters. The instrument used for this purpose was the Patient Dose Calibrator (PDC). It was developed to be used as a reference to check the calibration of PKA and air kerma meters that are used for dosimetry in patients and to verify the consistency and behavior of systems of automatic exposure control. Because it is a new equipment, which, in Brazil, is not yet used as reference equipment for calibration, it was also performed the quality control of this equipment with characterization tests, the calibration and an evaluation of the energy dependence. After the tests, it was proved that the PDC can be used as a reference instrument and that the calibration must be performed in situ, so that the characteristics of each X-ray equipment, where the PKA meters are used, are considered. The calibration was then performed with portable PKA meters and in an interventional radiology equipment that has a PKA meter fixed. The results were good and it was proved the need for calibration of these meters and the importance of in situ calibration with a reference meter. (author)

  14. Innovative methodology for intercomparison of radionuclide calibrators using short half-life in situ prepared radioactive sources

    International Nuclear Information System (INIS)

    Oliveira, P. A.; Santos, J. A. M.

    2014-01-01

    Purpose: An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as 99m Tc or most positron emission tomography radiopharmaceuticals. Methods: By evaluation of the resulting net optical density (netOD) using a standardized scanning method of irradiated Gafchromic XRQA2 film, a comparison of the netOD measurement with a previously determined calibration curve can be made and the difference between the tested radionuclide calibrator and a radionuclide calibrator used as reference device can be calculated. To estimate the total expected measurement uncertainties, a careful analysis of the methodology, for the case of 99m Tc, was performed: reproducibility determination, scanning conditions, and possible fadeout effects. Since every factor of the activity measurement procedure can influence the final result, the method also evaluates correct syringe positioning inside the radionuclide calibrator. Results: As an alternative to using a calibrated source sent to the surveyed site, which requires a relatively long half-life of the nuclide, or sending a portable calibrated radionuclide calibrator, the proposed method uses a source preparedin situ. An indirect activity determination is achieved by the irradiation of a radiochromic film using 99m Tc under strictly controlled conditions, and cumulated activity calculation from the initial activity and total irradiation time. The irradiated Gafchromic film and the irradiator, without the source, can then be sent to a National Metrology Institute for evaluation of the results. Conclusions: The methodology described in this paper showed to have a good potential for accurate (3%) radionuclide calibrators intercomparison studies for 99m Tc between Nuclear Medicine centers without source transfer and can easily be adapted to other short half-life radionuclides

  15. Update of Part 61 Impacts Analysis Methodology. Methodology report. Volume 1

    International Nuclear Information System (INIS)

    Oztunali, O.I.; Roles, G.W.

    1986-01-01

    Under contract to the US Nuclear Regulatory Commission, the Envirosphere Company has expanded and updated the impacts analysis methodology used during the development of the 10 CFR Part 61 rule to allow improved consideration of the costs and impacts of treatment and disposal of low-level waste that is close to or exceeds Class C concentrations. The modifications described in this report principally include: (1) an update of the low-level radioactive waste source term, (2) consideration of additional alternative disposal technologies, (3) expansion of the methodology used to calculate disposal costs, (4) consideration of an additional exposure pathway involving direct human contact with disposed waste due to a hypothetical drilling scenario, and (5) use of updated health physics analysis procedures (ICRP-30). Volume 1 of this report describes the calculational algorithms of the updated analysis methodology

  16. Update of Part 61 Impacts Analysis Methodology. Methodology report. Volume 1

    Energy Technology Data Exchange (ETDEWEB)

    Oztunali, O.I.; Roles, G.W.

    1986-01-01

    Under contract to the US Nuclear Regulatory Commission, the Envirosphere Company has expanded and updated the impacts analysis methodology used during the development of the 10 CFR Part 61 rule to allow improved consideration of the costs and impacts of treatment and disposal of low-level waste that is close to or exceeds Class C concentrations. The modifications described in this report principally include: (1) an update of the low-level radioactive waste source term, (2) consideration of additional alternative disposal technologies, (3) expansion of the methodology used to calculate disposal costs, (4) consideration of an additional exposure pathway involving direct human contact with disposed waste due to a hypothetical drilling scenario, and (5) use of updated health physics analysis procedures (ICRP-30). Volume 1 of this report describes the calculational algorithms of the updated analysis methodology.

  17. Methodology for calibration of ionization chambers for X-ray of low energy in absorbed dose to water

    International Nuclear Information System (INIS)

    Oliveira, C.T.; Vivolo, V.; Potiens, M.P.A.

    2015-01-01

    The beams of low energy X-ray (10 to 150 kV) are used in several places in the world to treat a wide variety of surface disorders, and between these malignancies. As in Brazil, at this moment, there is no calibration laboratory providing the control service or calibration of parallel plate ionization chambers, the aim of this project was to establish a methodology for calibration of this kind of ionization chambers at low energy X-ray beams in terms of absorbed dose to water using simulators in the LCI. (author)

  18. Drop size distribution measured by imaging: determination of the measurement volume by the calibration of the point spread function

    International Nuclear Information System (INIS)

    Fdida, Nicolas; Blaisot, Jean-Bernard

    2010-01-01

    Measurement of drop size distributions in a spray depends on the definition of the control volume for drop counting. For image-based techniques, this implies the definition of a depth-of-field (DOF) criterion. A sizing procedure based on an imaging model and associated with a calibration procedure is presented. Relations between image parameters and object properties are used to provide a measure of the size of the droplets, whatever the distance from the in-focus plane. A DOF criterion independent of the size of the drops and based on the determination of the width of the point spread function (PSF) is proposed. It allows to extend the measurement volume to defocused droplets and, due to the calibration of the PSF, to clearly define the depth of the measurement volume. Calibrated opaque discs, calibrated pinholes and an optical edge are used for this calibration. A comparison of the technique with a phase Doppler particle analyser and a laser diffraction granulometer is performed on an application to an industrial spray. Good agreement is found between the techniques when particular care is given to the sampling of droplets. The determination of the measurement volume is used to determine the drop concentration in the spray and the maximum drop concentration that imaging can support

  19. Calibration of a solid state nuclear track detector for the measurements of volumic activity of Radon

    International Nuclear Information System (INIS)

    HAKAM, O.K.; LFERDE, M.; BERRADA, M.

    1994-01-01

    Time - integrated measurements of environmental radiation activity are commonly carried out using solid state nuclear track detectors ( SSNTD ). These detectors should be calibrated of volumic activity of radon. This paper reports the results of experiments conducted to calibrate cellulose nitrate films LR - 115 type II used for measurements of volumic activity of radon in indoor air in dwellings and enclosed work areas in Morocco. Calibration measurements were made in laboratory using a calibration chamber and a radon source. The calibration chamber is a cylindric box ( 2613,6 cm sup 3)which we have manufactured of aluminium. The radon source is a natural sample rich of aluminium (17,29 + 0 ,12) Bq/g. The films are placed in detector holder with membrane and exposed inside the calibration chamber to varying concentrations of radon. Following the exposure, the films were chemically etched in sodium hydroxide (2,5 N) at 60 C for 120 minutes. The number of registered alpha particle tracks were counted with an optical microscope. In the used etching conditions, the removed mean thickness is in the order of 6 micro m. Therefore, we have normalized the track density to this value . We obtained a calibration factor of 0, 58 tracks . cm sup -2/ K Bq . h . m sup -3 . 1 tab.; 1 fig.; 2 refs. (author)

  20. Sandia software guidelines: Volume 5, Tools, techniques, and methodologies

    Energy Technology Data Exchange (ETDEWEB)

    1989-07-01

    This volume is one in a series of Sandia Software Guidelines intended for use in producing quality software within Sandia National Laboratories. This volume describes software tools and methodologies available to Sandia personnel for the development of software, and outlines techniques that have proven useful within the Laboratories and elsewhere. References and evaluations by Sandia personnel are included. 6 figs.

  1. Methodology for the digital calibration of analog circuits and systems with case studies

    CERN Document Server

    Pastre, Marc

    2006-01-01

    Methodology for the Digital Calibration of Analog Circuits and Systems shows how to relax the extreme design constraints in analog circuits, allowing the realization of high-precision systems even with low-performance components. A complete methodology is proposed, and three applications are detailed. To start with, an in-depth analysis of existing compensation techniques for analog circuit imperfections is carried out. The M/2+M sub-binary digital-to-analog converter is thoroughly studied, and the use of this very low-area circuit in conjunction with a successive approximations algorithm for digital compensation is described. A complete methodology based on this compensation circuit and algorithm is then proposed. The detection and correction of analog circuit imperfections is studied, and a simulation tool allowing the transparent simulation of analog circuits with automatic compensation blocks is introduced. The first application shows how the sub-binary M/2+M structure can be employed as a conventional di...

  2. Contribution to the RMTC in the field of tank calibration and measurements - the TAMSCA laboratory

    International Nuclear Information System (INIS)

    Hunt, B.A.; Landat, D.; Caviglia, M.; Silvapestana, L.

    1999-01-01

    The Russian Methodological and Training Centre (RMTC) is being established for training of personnel from the various Russian and CIS nuclear facilities organizations in the control and accountancy methods, utilised in EURATOM and in the IAEA. Under the project equipment and support will be provided in a number of areas, namely containment and surveillance, training, passive/active neutron assay and mass/volume methodologies. For the latter a mass/volume measurement laboratory - a Tank Measurements and Calibration Laboratory (TAMSCA) is being set-up in IPPE, Obninsk. The goal is to upgrade the methodology within the Russian Federation in the application of mass/volume measurement techniques and render a facility suitable adapted to carrying out training courses with specific orientation for the nuclear inspectors and operators of nuclear facilities for nuclear accountancy and control [ru

  3. Efficient Calibration of Distributed Catchment Models Using Perceptual Understanding and Hydrologic Signatures

    Science.gov (United States)

    Hutton, C.; Wagener, T.; Freer, J. E.; Duffy, C.; Han, D.

    2015-12-01

    Distributed models offer the potential to resolve catchment systems in more detail, and therefore simulate the hydrological impacts of spatial changes in catchment forcing (e.g. landscape change). Such models may contain a large number of model parameters which are computationally expensive to calibrate. Even when calibration is possible, insufficient data can result in model parameter and structural equifinality. In order to help reduce the space of feasible models and supplement traditional outlet discharge calibration data, semi-quantitative information (e.g. knowledge of relative groundwater levels), may also be used to identify behavioural models when applied to constrain spatially distributed predictions of states and fluxes. The challenge is to combine these different sources of information together to identify a behavioural region of state-space, and efficiently search a large, complex parameter space to identify behavioural parameter sets that produce predictions that fall within this behavioural region. Here we present a methodology to incorporate different sources of data to efficiently calibrate distributed catchment models. Metrics of model performance may be derived from multiple sources of data (e.g. perceptual understanding and measured or regionalised hydrologic signatures). For each metric, an interval or inequality is used to define the behaviour of the catchment system, accounting for data uncertainties. These intervals are then combined to produce a hyper-volume in state space. The state space is then recast as a multi-objective optimisation problem, and the Borg MOEA is applied to first find, and then populate the hyper-volume, thereby identifying acceptable model parameter sets. We apply the methodology to calibrate the PIHM model at Plynlimon, UK by incorporating perceptual and hydrologic data into the calibration problem. Furthermore, we explore how to improve calibration efficiency through search initialisation from shorter model runs.

  4. Comparison of calibration strategies for optical 3D scanners based on structured light projection using a new evaluation methodology

    Science.gov (United States)

    Bräuer-Burchardt, Christian; Ölsner, Sandy; Kühmstedt, Peter; Notni, Gunther

    2017-06-01

    In this paper a new evaluation strategy for optical 3D scanners based on structured light projection is introduced. It can be used for the characterization of the expected measurement accuracy. Compared to the procedure proposed in the VDI/VDE guidelines for optical 3D measurement systems based on area scanning it requires less effort and provides more impartiality. The methodology is suitable for the evaluation of sets of calibration parameters, which mainly determine the quality of the measurement result. It was applied to several calibrations of a mobile stereo camera based optical 3D scanner. The performed calibrations followed different strategies regarding calibration bodies and arrangement of the observed scene. The results obtained by the different calibration strategies are discussed and suggestions concerning future work on this area are given.

  5. A proposed methodology for computational fluid dynamics code verification, calibration, and validation

    Science.gov (United States)

    Aeschliman, D. P.; Oberkampf, W. L.; Blottner, F. G.

    Verification, calibration, and validation (VCV) of Computational Fluid Dynamics (CFD) codes is an essential element of the code development process. The exact manner in which code VCV activities are planned and conducted, however, is critically important. It is suggested that the way in which code validation, in particular, is often conducted--by comparison to published experimental data obtained for other purposes--is in general difficult and unsatisfactory, and that a different approach is required. This paper describes a proposed methodology for CFD code VCV that meets the technical requirements and is philosophically consistent with code development needs. The proposed methodology stresses teamwork and cooperation between code developers and experimentalists throughout the VCV process, and takes advantage of certain synergisms between CFD and experiment. A novel approach to uncertainty analysis is described which can both distinguish between and quantify various types of experimental error, and whose attributes are used to help define an appropriate experimental design for code VCV experiments. The methodology is demonstrated with an example of laminar, hypersonic, near perfect gas, 3-dimensional flow over a sliced sphere/cone of varying geometrical complexity.

  6. Employing an Incentive Spirometer to Calibrate Tidal Volumes Estimated from a Smartphone Camera

    Science.gov (United States)

    Reyes, Bersain A.; Reljin, Natasa; Kong, Youngsun; Nam, Yunyoung; Ha, Sangho; Chon, Ki H.

    2016-01-01

    A smartphone-based tidal volume (VT) estimator was recently introduced by our research group, where an Android application provides a chest movement signal whose peak-to-peak amplitude is highly correlated with reference VT measured by a spirometer. We found a Normalized Root Mean Squared Error (NRMSE) of 14.998% ± 5.171% (mean ± SD) when the smartphone measures were calibrated using spirometer data. However, the availability of a spirometer device for calibration is not realistic outside clinical or research environments. In order to be used by the general population on a daily basis, a simple calibration procedure not relying on specialized devices is required. In this study, we propose taking advantage of the linear correlation between smartphone measurements and VT to obtain a calibration model using information computed while the subject breathes through a commercially-available incentive spirometer (IS). Experiments were performed on twelve (N = 12) healthy subjects. In addition to corroborating findings from our previous study using a spirometer for calibration, we found that the calibration procedure using an IS resulted in a fixed bias of −0.051 L and a RMSE of 0.189 ± 0.074 L corresponding to 18.559% ± 6.579% when normalized. Although it has a small underestimation and slightly increased error, the proposed calibration procedure using an IS has the advantages of being simple, fast, and affordable. This study supports the feasibility of developing a portable smartphone-based breathing status monitor that provides information about breathing depth, in addition to the more commonly estimated respiratory rate, on a daily basis. PMID:26999152

  7. Employing an Incentive Spirometer to Calibrate Tidal Volumes Estimated from a Smartphone Camera

    Directory of Open Access Journals (Sweden)

    Bersain A. Reyes

    2016-03-01

    Full Text Available A smartphone-based tidal volume (VT estimator was recently introduced by our research group, where an Android application provides a chest movement signal whose peak-to-peak amplitude is highly correlated with reference VT measured by a spirometer. We found a Normalized Root Mean Squared Error (NRMSE of 14.998% ± 5.171% (mean ± SD when the smartphone measures were calibrated using spirometer data. However, the availability of a spirometer device for calibration is not realistic outside clinical or research environments. In order to be used by the general population on a daily basis, a simple calibration procedure not relying on specialized devices is required. In this study, we propose taking advantage of the linear correlation between smartphone measurements and VT to obtain a calibration model using information computed while the subject breathes through a commercially-available incentive spirometer (IS. Experiments were performed on twelve (N = 12 healthy subjects. In addition to corroborating findings from our previous study using a spirometer for calibration, we found that the calibration procedure using an IS resulted in a fixed bias of −0.051 L and a RMSE of 0.189 ± 0.074 L corresponding to 18.559% ± 6.579% when normalized. Although it has a small underestimation and slightly increased error, the proposed calibration procedure using an IS has the advantages of being simple, fast, and affordable. This study supports the feasibility of developing a portable smartphone-based breathing status monitor that provides information about breathing depth, in addition to the more commonly estimated respiratory rate, on a daily basis.

  8. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 4: Accurate determination of liquid height in accountancy tanks equipped with dip tubes, slow bubbling rate

    International Nuclear Information System (INIS)

    2008-01-01

    ISO 18213 deals with the acquisition, standardization, analysis, and use of calibration to determine liquid volumes in process tanks for the accountancy of nuclear materials. This part of ISO 18213 is complementary to the other parts, ISO 18213-1 (procedural overview), ISO 18213-2 (data standardization), ISO 18213-3 (statistical methods), ISO 18213-5 (fast bubbling rate) and ISO 18213-6 (in-tank determination of liquid density). The procedure presented herein for determining liquid height from measurements of induced pressure applies specifically when a very slow bubbling rate is employed. A similar procedure that is appropriate for a fast bubbling rate is given in ISO 18213-5. Measurements of the volume and height of liquid in a process accountancy tank are often made in order to estimate or verify the tank's calibration or volume measurement equation. The calibration equation relates the response of the tank's measurement system to some independent measure of tank volume. Beginning with an empty tank, calibration data are typically acquired by introducing a series of carefully measured quantities of some calibration liquid into the tank. The quantity of liquid added, the response of the tank's measurement system, and relevant ambient conditions such as temperature are measured for each incremental addition. Several calibration runs are made to obtain data for estimating or verifying a tank's calibration or measurement equation. A procedural overview of the tank calibration and volume measurement process is given in ISO 18213-1. An algorithm for standardizing tank calibration and volume measurement data to minimize the effects of variability in ambient conditions that prevail during the measurement period is given in ISO 18213-2. The procedure presented in this part of ISO 18213 for determining the height of calibration liquid in the tank from a measurement of the pressure it induces in the tank's measurement system is a vital component of that algorithm. In some

  9. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 5: Accurate determination of liquid height in accountancy tanks equipped with dip tubes, fast bubbling rate

    International Nuclear Information System (INIS)

    2008-01-01

    ISO 18213 deals with the acquisition, standardization, analysis, and use of calibration to determine liquid volumes in process tanks for the accountancy of nuclear materials. This part of ISO 18213 is complementary to the other parts, ISO 18213-1 (procedural overview), ISO 18213-2 (data standardization), ISO 18213-3 (statistical methods), ISO 18213-5 (fast bubbling rate) and ISO 18213-6 (in-tank determination of liquid density). The procedure presented herein for determining liquid height from measurements of induced pressure applies specifically when a very slow bubbling rate is employed. A similar procedure that is appropriate for a fast bubbling rate is given in ISO 18213-5. Measurements of the volume and height of liquid in a process accountancy tank are often made in order to estimate or verify the tank's calibration or volume measurement equation. The calibration equation relates the response of the tank's measurement system to some independent measure of tank volume. Beginning with an empty tank, calibration data are typically acquired by introducing a series of carefully measured quantities of some calibration liquid into the tank. The quantity of liquid added, the response of the tank's measurement system, and relevant ambient conditions such as temperature are measured for each incremental addition. Several calibration runs are made to obtain data for estimating or verifying a tank's calibration or measurement equation. A procedural overview of the tank calibration and volume measurement process is given in ISO 18213-1. An algorithm for standardizing tank calibration and volume measurement data to minimize the effects of variability in ambient conditions that prevail during the measurement period is given in ISO 18213-2. The procedure presented in this part of ISO 18213 for determining the height of calibration liquid in the tank from a measurement of the pressure it induces in the tank's measurement system is a vital component of that algorithm. In some

  10. Calibration of CORSIM models under saturated traffic flow conditions.

    Science.gov (United States)

    2013-09-01

    This study proposes a methodology to calibrate microscopic traffic flow simulation models. : The proposed methodology has the capability to calibrate simultaneously all the calibration : parameters as well as demand patterns for any network topology....

  11. Evaluation of a laser scanner for large volume coordinate metrology: a comparison of results before and after factory calibration

    International Nuclear Information System (INIS)

    Ferrucci, M; Muralikrishnan, B; Sawyer, D; Phillips, S; Petrov, P; Yakovlev, Y; Astrelin, A; Milligan, S; Palmateer, J

    2014-01-01

    Large volume laser scanners are increasingly being used for a variety of dimensional metrology applications. Methods to evaluate the performance of these scanners are still under development and there are currently no documentary standards available. This paper describes the results of extensive ranging and volumetric performance tests conducted on a large volume laser scanner. The results demonstrated small but clear systematic errors that are explained in the context of a geometric error model for the instrument. The instrument was subsequently returned to the manufacturer for factory calibration. The ranging and volumetric tests were performed again and the results are compared against those obtained prior to the factory calibration. (paper)

  12. The peak efficiency calibration of volume source using 152Eu point source in computer

    International Nuclear Information System (INIS)

    Shen Tingyun; Qian Jianfu; Nan Qinliang; Zhou Yanguo

    1997-01-01

    The author describes the method of the peak efficiency calibration of volume source by means of 152 Eu point source for HPGe γ spectrometer. The peak efficiency can be computed by Monte Carlo simulation, after inputting parameter of detector. The computation results are in agreement with the experimental results with an error of +-3.8%, with an exception one is about +-7.4%

  13. Semi-empirical approach for calibration of CR-39 detectors in diffusion chambers for radon measurements

    International Nuclear Information System (INIS)

    Pereyra A, P.; Lopez H, M. E.; Palacios F, D.; Sajo B, L.; Valdivia, P.

    2016-10-01

    Simulated and measured calibration of PADC detectors is given for cylindrical diffusion chambers employed in environmental radon measurements. The method is based on determining the minimum alpha energy (E min ), average critical angle (<Θ c >), and fraction of 218 Po atoms; the volume of the chamber (f 1 ), are compared to commercially available devices. Radon concentration for exposed detectors is obtained from induced track densities and the well-established calibration coefficient for NRPB monitor. Calibration coefficient of a PADC detector in a cylindrical diffusion chamber of any size is determined under the same chemical etching conditions and track analysis methodology. In this study the results of numerical examples and comparison between experimental calibration coefficients and simulation purpose made code. Results show that the developed method is applicable when uncertainties of 10% are acceptable. (Author)

  14. A rational methodology for the study of foundations for marine structures

    International Nuclear Information System (INIS)

    Mira Mc Willams, P.; Fernandez-Merodo, J. A.; Pastor Perez, M.; Monte Saez, J. L.; Martinez Santamaria, J. M.; Cuellar Mirasol, V.; Martin Baanante, M. E.; Rodriguez Sanchez-Arevalo, I; Lopez Maldonando, J. D.; Tomas Sampedro, A.

    2011-01-01

    A methodology for the study of marine foundations is presented. The response in displacements, stresses and pore water pressures in obtained from a finite element coupled formulation. Loads due to wave action of the foundation are obtained from a volume of fluid type fluid-structure interaction numerical model. Additionally, the methodology includes a Generalized Plasticity based constitutive model for granular materials capable of representing liquefaction fenomena of sands subjected to cyclic loading, such as those frequently appearing in the problems studied. Calibration of this model requires a series of laboratory tests detailed herein. This methodology is applied to the study of the response of a caisson breakwater foundation. (Author) 10 refs.

  15. Design, calibration and error analysis of instrumentation for heat transfer measurements in internal combustion engines

    Science.gov (United States)

    Ferguson, C. R.; Tree, D. R.; Dewitt, D. P.; Wahiduzzaman, S. A. H.

    1987-01-01

    The paper reports the methodology and uncertainty analyses of instrumentation for heat transfer measurements in internal combustion engines. Results are presented for determining the local wall heat flux in an internal combustion engine (using a surface thermocouple-type heat flux gage) and the apparent flame-temperature and soot volume fraction path length product in a diesel engine (using two-color pyrometry). It is shown that a surface thermocouple heat transfer gage suitably constructed and calibrated will have an accuracy of 5 to 10 percent. It is also shown that, when applying two-color pyrometry to measure the apparent flame temperature and soot volume fraction-path length, it is important to choose at least one of the two wavelengths to lie in the range of 1.3 to 2.3 micrometers. Carefully calibrated two-color pyrometer can ensure that random errors in the apparent flame temperature and in the soot volume fraction path length will remain small (within about 1 percent and 10-percent, respectively).

  16. Calibration of Ge(Li) semiconductor detector by method using agar volume source

    International Nuclear Information System (INIS)

    Yanase, Nobuyuki; Kasai, Atsushi

    1979-12-01

    The Ge(Li) semiconductor detector was calibrated for measurements of environmental samples. The radioisotopes used for standard sources are 22 Na, 51 Cr, 56 Co, 57 Co, 133 Ba, 137 Cs, 144 Ce and 241 Am. These are mixed with hot agar aqueous solution and fixed uniformly in a cylindrical plastic case in cooling. The agar volume source is advantageous in handling over the fluid aqueous source. The prepared cylindrical standard sources are in diameters 6 and 8 cm and thicknesses 1, 5, 10, 20, 30 and 40 mm (only for 8 cm diameter). The radioactivities of prepared standard sources are between 0.03 μCi and 0.2 μCi. It takes only a week to make the calibration except data processing. The obtained full energy peak efficiency curves include 5 - 10% error due to preparation of agar source, reference radioactivity data of purchased standard solutions, reference data of branching ratio of gamma-ray and sum effect. The efficiency curves, however, are sufficient for quantitative analysis of environmental samples. (author)

  17. 2D-3D Registration of CT Vertebra Volume to Fluoroscopy Projection: A Calibration Model Assessment

    Directory of Open Access Journals (Sweden)

    P. Bifulco

    2010-01-01

    Full Text Available This study extends a previous research concerning intervertebral motion registration by means of 2D dynamic fluoroscopy to obtain a more comprehensive 3D description of vertebral kinematics. The problem of estimating the 3D rigid pose of a CT volume of a vertebra from its 2D X-ray fluoroscopy projection is addressed. 2D-3D registration is obtained maximising a measure of similarity between Digitally Reconstructed Radiographs (obtained from the CT volume and real fluoroscopic projection. X-ray energy correction was performed. To assess the method a calibration model was realised a sheep dry vertebra was rigidly fixed to a frame of reference including metallic markers. Accurate measurement of 3D orientation was obtained via single-camera calibration of the markers and held as true 3D vertebra position; then, vertebra 3D pose was estimated and results compared. Error analysis revealed accuracy of the order of 0.1 degree for the rotation angles of about 1 mm for displacements parallel to the fluoroscopic plane, and of order of 10 mm for the orthogonal displacement.

  18. Semi-empirical approach for calibration of CR-39 detectors in diffusion chambers for radon measurements

    Energy Technology Data Exchange (ETDEWEB)

    Pereyra A, P.; Lopez H, M. E. [Pontificia Universidad Catolica del Peru, Av. Universitaria 1801, San Miguel Lima 32 (Peru); Palacios F, D.; Sajo B, L. [Universidad Simon Bolivar, Laboratorio de Fisica Nuclear, Apartado 89000 Caracas (Venezuela, Bolivarian Republic of); Valdivia, P., E-mail: ppereyr@pucp.edu.pe [Universidad Nacional de Ingenieria, Av. Tupac Amaru s/n, Rimac, Lima 25 (Peru)

    2016-10-15

    Simulated and measured calibration of PADC detectors is given for cylindrical diffusion chambers employed in environmental radon measurements. The method is based on determining the minimum alpha energy (E{sub min}), average critical angle (<Θ{sub c}>), and fraction of {sup 218}Po atoms; the volume of the chamber (f{sub 1}), are compared to commercially available devices. Radon concentration for exposed detectors is obtained from induced track densities and the well-established calibration coefficient for NRPB monitor. Calibration coefficient of a PADC detector in a cylindrical diffusion chamber of any size is determined under the same chemical etching conditions and track analysis methodology. In this study the results of numerical examples and comparison between experimental calibration coefficients and simulation purpose made code. Results show that the developed method is applicable when uncertainties of 10% are acceptable. (Author)

  19. Radioactivity measurement of 18F in 16 ml vials for calibration of radionuclide calibrators

    International Nuclear Information System (INIS)

    Wurdiyanto, Gatot; Marsoem, Pujadi; Candra, Hermawan; Wijono, Paidi

    2012-01-01

    Fluorine-18 is obtained through the reaction 18 O(p, n) 18 F using a cyclotron that is situated in a hospital in Jakarta. Standardization of the 18 F solution is performed by gamma spectrometry using calibration sources of 152 Eu, 60 Co and 137 Cs that have traceability to the International System of units (SI). The activities in the 16 ml vials that were used for calibrating the radionuclide calibrators were between 1 and 2 GBq, with expanded uncertainties of 3.8%. The expanded uncertainty, at a coverage factor of k=2, on the derived calibration factor for the radionuclide calibrator was 6.6%. - Highlights: ► PTKMR–BATAN as a NMI of Indonesia is required to have procedures to calibrate the radionuclide calibrators. ► Standardizations were carried out on a solution of [ 18 F]FDG using gamma spectrometry. ► The volume of 18 F solutions used was 16 ml because this is the volume often used in hospitals. ► The Secondary Standard ionization chamber is a CRC-7BT Capintec radionuclide calibrator. ► A dial setting for 16 ml of [ 18 F]FDG solution in a vial is 443 for the Capintec dose calibrator.

  20. Waste isolation in the U.S., technical programs and public education. Volume 2 - low level waste, volume reduction methodologies and economics. Vol. 2

    International Nuclear Information System (INIS)

    Post, R.G.

    1984-01-01

    This volume presents information regarding low-level waste, volume reduction methodologies and economics. Topics include: public education on nuclear waste; economics of low-level waste management systems; operating experience with advanced volume reduction techniques; solidification of waste; operating experience with advanced volume reduction techniques--incineration; regional plans for the disposal of low-level waste; radwaste system modifications at nuclear power plants; operating experience with advanced volume reduction techniques--operations and on-site storage issues; and economic impact of 10CFR61

  1. Optimization of SPECT calibration for quantification of images applied to dosimetry with iodine-131

    International Nuclear Information System (INIS)

    Carvalho, Samira Marques de

    2018-01-01

    SPECT systems calibration plays an essential role in the accuracy of the quantification of images. In this work, in its first stage, an optimized SPECT calibration method was proposed for 131 I studies, considering the partial volume effect (PVE) and the position of the calibration source. In the second stage, the study aimed to investigate the impact of count density and reconstruction parameters on the determination of the calibration factor and the quantification of the image in dosimetry studies, considering the reality of clinical practice in Brazil. In the final step, the study aimed evaluating the influence of several factors in the calibration for absorbed dose calculation using Monte Carlo simulations (MC) GATE code. Calibration was performed by determining a calibration curve (sensitivity versus volume) obtained by applying different thresholds. Then, the calibration factors were determined with an exponential function adjustment. Images were performed with high and low counts densities for several source positions within the simulator. To validate the calibration method, the calibration factors were used for absolute quantification of the total reference activities. The images were reconstructed adopting two approaches of different parameters, usually used in patient images. The methodology developed for the calibration of the tomographic system was easier and faster to implement than other procedures suggested to improve the accuracy of the results. The study also revealed the influence of the location of the calibration source, demonstrating better precision in the absolute quantification considering the location of the target region during the calibration of the system. The study applied in the Brazilian thyroid protocol suggests the revision of the calibration of the SPECT system, including different positions for the reference source, besides acquisitions considering the Signal to Noise Ratio (SNR) of the images. Finally, the doses obtained with the

  2. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  3. Clouds and the Earth's Radiant Energy System (CERES) algorithm theoretical basis document. volume 2; Geolocation, calibration, and ERBE-like analyses (subsystems 1-3)

    Science.gov (United States)

    Wielicki, B. A. (Principal Investigator); Barkstrom, B. R. (Principal Investigator); Charlock, T. P.; Baum, B. A.; Green, R. N.; Minnis, P.; Smith, G. L.; Coakley, J. A.; Randall, D. R.; Lee, R. B., III

    1995-01-01

    The theoretical bases for the Release 1 algorithms that will be used to process satellite data for investigation of the Clouds and Earth's Radiant Energy System (CERES) are described. The architecture for software implementation of the methodologies is outlined. Volume 2 details the techniques used to geolocate and calibrate the CERES scanning radiometer measurements of shortwave and longwave radiance to invert the radiances to top-of-the-atmosphere (TOA) and surface fluxes following the Earth Radiation Budget Experiment (ERBE) approach, and to average the fluxes over various time and spatial scales to produce an ERBE-like product. Spacecraft ephemeris and sensor telemetry are used with calibration coefficients to produce a chronologically ordered data product called bidirectional scan (BDS) radiances. A spatially organized instrument Earth scan product is developed for the cloud-processing subsystem. The ERBE-like inversion subsystem converts BDS radiances to unfiltered instantaneous TOA and surface fluxes. The TOA fluxes are determined by using established ERBE techniques. Hourly TOA fluxes are computed from the instantaneous values by using ERBE methods. Hourly surface fluxes are estimated from TOA fluxes by using simple parameterizations based on recent research. The averaging process produces daily, monthly-hourly, and monthly means of TOA and surface fluxes at various scales. This product provides a continuation of the ERBE record.

  4. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten

    2007-01-01

    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve...... the statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited...

  5. Impacts of Outer Continental Shelf (OCS) development on recreation and tourism. Volume 3. Detailed methodology

    Energy Technology Data Exchange (ETDEWEB)

    1987-04-01

    The final report for the project is presented in five volumes. This volume, Detailed Methodology Review, presents a discussion of the methods considered and used to estimate the impacts of Outer Continental Shelf (OCS) oil and gas development on coastal recreation in California. The purpose is to provide the Minerals Management Service with data and methods to improve their ability to analyze the socio-economic impacts of OCS development. Chapter II provides a review of previous attempts to evaluate the effects of OCS development and of oil spills on coastal recreation. The review also discusses the strengths and weaknesses of different approaches and presents the rationale for the methodology selection made. Chapter III presents a detailed discussion of the methods actually used in the study. The volume contains the bibliography for the entire study.

  6. Calibration of alpha surface contamination monitor

    International Nuclear Information System (INIS)

    Freitas, I.S.M. de; Goncalez, O.L.

    1990-01-01

    In this work, the results, as well as the methodology, of the calibration of an alpha surface contamination monitor are presented. The calibration factors are obtained by least-squares fitting with effective variance. (author)

  7. Development of a calibration methodology for instruments used to interventional radiology quality control

    International Nuclear Information System (INIS)

    Miranda, Jurema Aparecida de

    2009-01-01

    Interventional radiology is the technique where X radiation images are used as a tool in the conduction of diagnostic or/and therapeutic procedures. The exposition times are long for both procedures, diagnostic and therapeutic, may cause serious injuries in the patient, and also contribute to the dose of the clinical staff. In Brazil there are not yet well established rules to determine the doses and to make the dosimetry in fluoroscopic beams. There is great interest in this study, in relation to the beam quality, the half-value-layer, and others parameters. In this work a Medicor Neo Diagnomax clinical X ray generator, fluoroscopy mode, was used to develop a calibration methodology for instruments used in interventional radiology quality control. One plane parallel ionization chamber PTW was used as monitor. The ionization chambers recommended for fluoroscopy measurements had been evaluated and calibrated in relation to the IPEN Calibration Laboratory reference ionization chamber. The RQR3, RQR5 and RQR7 radiation qualities and the specific ones for fluoroscopy, RQC3, RQC5 and RQC7, were established following the norm IEC 61267. All beams characteristics were determined. Ionization chambers positioning system and the acrylic phantoms to the entrance and exit doses determination were developed and constructed. The results obtained show air kerma rates of 4.5x10 -3 , 1.2x10 -2 and 1.9x10 -2 Gy/min for RQC3, RQC5 and RQC7 respectively. Tests with and without the collimation just after the monitor chamber, were carried out and the results showed a difference of +5.5%, +0.6% e + 0.8%, confirming the importance of the collimation use in these interventionist procedures. (author)

  8. The Observability Calibration Test Development Framework

    Energy Technology Data Exchange (ETDEWEB)

    Endicott-Popovsky, Barbara E.; Frincke, Deborah A.

    2007-06-20

    Abstract— Formal standards, precedents, and best practices for verifying and validating the behavior of low layer network devices used for digital evidence-collection on networks are badly needed— initially so that these can be employed directly by device owners and data users to document the behaviors of these devices for courtroom presentation, and ultimately so that calibration testing and calibration regimes are established and standardized as common practice for both vendors and their customers [1]. The ultimate intent is to achieve a state of confidence in device calibration that allows the network data gathered by them to be relied upon by all parties in a court of law. This paper describes a methodology for calibrating forensic-ready low layer network devices based on the Flaw Hypothesis Methodology [2,3].

  9. Calibration of detectors type CR-39 for methodology implementation for Radon-222 determination in CRCN-NE, Brazil

    International Nuclear Information System (INIS)

    Silva, Karolayne E.M. da; Santos, Mariana L. de O.; Amaral, Déric S. do; Vilela, Eldice C.; França, Elvis J. de; Hazin, Clovis A.; Farias, Emerson E.G. de

    2017-01-01

    Radon-222 is a radioactive gas, a product of the decay of uranium-238, which emits alpha particles and represents more than 50% of the dose of natural radiation received by the population. Therefore, monitoring of this gas is essential. For indoor measurement, solid state detectors can be used, the most common of which is CR-39. For monitoring using CR-39, alpha particles, generated by radon-222 and the daughter radionuclides, strike the surface of the detector and generate traces. To relate the trace density per exposure area in environments with unknown activity concentration, it is necessary to determine the calibration factor. The objective of this study was to calibrate CR-39 type detectors for the implementation of the radon determination methodology in Centro Regional de Ciencias Nucleares do Nordeste - CRCN-NE of Brazilian Nuclear Energy Commission - CNEN. In order to determine the CR-39 calibration factor, 19 exposures of the detectors were performed in the CRCN-NE calibration chamber (RN1-CRCN) at an activity of 5.00 kBq m -3 , with the exposure time varying from 24 to 850 hours. For the detection of the detectors, sodium hydroxide was used in a thermostat bath at 90 ° C for 5 hours. The count of number of traits per unit of field was performed with the aid of optical microscopy with an increase of 100 times, being read 30 fields per dosimeters. As a result, the calibration factor was obtained, and the linear response of the trace density as a function of exposure was observed. The results allow the use of CR-39 in the determination of radon-222 by CRCN-NE

  10. Methodology for calibration of detector of Nal (TI) 3 'X 3' for measurements in vivo of patients with hyperthyroidism undergoing radioiodine therapy

    International Nuclear Information System (INIS)

    Carvalho, Carlaine B.; Lacerda, Isabelle V.B.; Oliveira, Mercia L.; Hazin, Clovis A.; Lima, Fabiana F.

    2013-01-01

    The aim of this study is to establish the methodology for calibration of the detection system to be used in determining the therapeutic activity of 131 I required to release the desired absorbed dose in the thyroid. This step is critical to the development of a protocol for individualized doses. The system consists of a detector of NaI (Tl ) 3 'x3' coupled to Genie 2000 software. The calibration sources of 60 CO, 137 Cs and 133 Ba were used. Obtained straight calibration system, with 60 CO and 137 Cs sources. Subsequently, the detector was calibrated using a simulator -neck thyroid designed and produced by the IRD/CNEN with known standard solution containing 18.7 kBq 133 Ba activity (in 12/09/24) evenly distributed. He was also calibrated with other thyroid - neck phantom Model 3108 manufactured by Searle Radigraphics Ind., containing a net source of 131 I ( 7.7 MBq ). Five measurements of five minutes were realized for three different distances detector simulator, and the respective calculated calibration factors was performed to three. The values of the calibration factors found for the simulator manufactured by IRD and the Searle Radigraphics Ind. for the distances 20 , 25 and 30cm were 0.35, 0.24, 0.18, and 0.15, 0.11, 0.09, respectively. With the detection system properly calibrated and the calibration factors established, the technique is suitable for the evaluation of diagnostic activity of 131 I incorporated by hyperthyroidism

  11. 40 CFR 86.519-90 - Constant volume sampler calibration.

    Science.gov (United States)

    2010-07-01

    ... total calibration. Allow the system to stabilize for 3 minutes and repeat the data acquisition. (7) Data... decrease as reflected by lower values for M. Calibrations should be performed at pump startup and after...

  12. Novel gravimetric measurement technique for quantitative volume calibration in the sub-microliter range

    International Nuclear Information System (INIS)

    Liang, Dong; Zengerle, Roland; Steinert, Chris; Ernst, Andreas; Koltay, Peter; Bammesberger, Stefan; Tanguy, Laurent

    2013-01-01

    We present a novel measurement method based on the gravimetric principles adapted from the ASTM E542 and ISO 4787 standards for quantitative volume determination in the sub-microliter range. Such a method is particularly important for the calibration of non-contact micro dispensers as well as other microfluidic devices. The novel method is based on the linear regression analysis of continuously monitored gravimetric results and therefore is referred to as ‘gravimetric regression method (GRM)’. In this context, the regression analysis is necessary to compensate the mass loss due to evaporation that is significant for very small dispensing volumes. A full assessment of the measurement uncertainty of GRM is presented and results in a standard measurement uncertainty around 6 nl for dosage volumes in the range from 40 nl to 1 µl. The GRM has been experimentally benchmarked with a dual-dye ratiometric photometric method (Artel Inc., Westbrook, ME, USA), which can provide traceability of measurement to the International System of Units (SI) through reference standards maintained by NIST. Good precision (max. CV = 2.8%) and consistency (bias around 7 nl in the volume range from 40 to 400 nl) have been observed comparing the two methods. Based on the ASTM and ISO standards on the one hand and the benchmark with the photometric method on the other hand, two different approaches for establishing traceability for the GRM are discussed. (paper)

  13. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities

    International Nuclear Information System (INIS)

    Costa, Alessandro Martins da

    1999-01-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  14. Methodology for calibration of detector of NaI (TI)) 3 ' X 3 ' for in vivo measurements of patients with hyperthyroidism undergoing to radioiodotherapy

    International Nuclear Information System (INIS)

    Carvalho, Carlaine B.; Lacerda, Isabelle V.B.; Oliveira, Mercia L.; Hazin, Clovis A.; Lima, Fabiana F.

    2013-01-01

    The aim of this study is to establish the methodology for calibration of the detection system to be used in determining the therapeutic activity of 131 I required to release desired absorbed dose in the thyroid gland . This step is critical to the development of a protocol for individualized doses. The system consists of a detector of NaI (Tl ) 3'x3' coupled to software Genie 2000. We used the calibration sources of 60 Co , 137 Cs and 133 Ba. We obtained the straight calibration system, with sources 60 Co and 137 Cs. Subsequently , the detector was calibrated using a thyroid phantom-neck designed and produced by the IRD / CNEN with known activity of 133 Ba standard solution containing 18.7 kBq (on 09/24/12) evenly distributed. He was also calibrated with other thyroid- neck phantom model 3108 manufactured by Searle Radigraphics Ind., containing a liquid source of 131 I ( 7.7 MBq ). Five measurements were performed during 5 minutes for three different distances detector-simulator and calculated the corresponding calibration factors . The values of the calibration factors found for the simulator made by IRD and Searle Radigraphics Ind. for the distances 20, 25 and 30 cm were 0.35 , 0.24, 0.18, 0.15 , 0.11, 0, 09 , respectively. With the detection system properly calibrated and the calibration factors established, the technique is suitable for the evaluation of diagnostic activities of 131 I incorporated by hyperthyroid patients. (author)

  15. Compact radiometric microwave calibrator

    International Nuclear Information System (INIS)

    Fixsen, D. J.; Wollack, E. J.; Kogut, A.; Limon, M.; Mirel, P.; Singal, J.; Fixsen, S. M.

    2006-01-01

    The calibration methods for the ARCADE II instrument are described and the accuracy estimated. The Steelcast coated aluminum cones which comprise the calibrator have a low reflection while maintaining 94% of the absorber volume within 5 mK of the base temperature (modeled). The calibrator demonstrates an absorber with the active part less than one wavelength thick and only marginally larger than the mouth of the largest horn and yet black (less than -40 dB or 0.01% reflection) over five octaves in frequency

  16. SeaWiFS Postlaunch Calibration and Validation Analyses

    Science.gov (United States)

    Hooker, Stanford B. (Editor); Firestone, Elaine (Editor); McClain, Charles R.; Barnes, Robert A.; Eplee, Robert E., Jr.; Franz, Bryan A.; Hsu, N. Christina; Patt, Frederick S.; Pietras, Christophe M.; Robinson, Wayne D.

    2000-01-01

    The effort to resolve data quality issues and improve on the initial data evaluation methodologies of the SeaWiFS Project was an extensive one. These evaluations have resulted, to date, in three major reprocessings of the entire data set where each reprocessing addressed the data quality issues that could be identified up to the time of the reprocessing. Three volumes of the SeaWiFS Postlaunch Technical Report Series (Volumes 9, 10, and 11) are needed to document the improvements implemented since launch. Volume 10 continues the sequential presentation of postlaunch data analysis and algorithm descriptions begun in Volume 9. Chapter 1 of Volume 10 describes an absorbing aerosol index, similar to that produced by the Total Ozone Mapping Spectrometer (TOMS) Project, which is used to flag pixels contaminated by absorbing aerosols, such as, dust and smoke. Chapter 2 discusses the algorithm being used to remove SeaWiFS out-of-band radiance from the water-leaving radiances. Chapter 3 provides an itemization of all significant changes in the processing algorithms for each of the first three reprocessings. Chapter 4 shows the time series of global clear water and deep-water (depths greater than 1,000m) bio-optical and atmospheric properties (normalized water-leaving radiances, chlorophyll, atmospheric optical depth, etc.) based on the eight-day composites as a check on the sensor calibration stability. Chapter 5 examines the variation in the derived products with scan angle using high resolution data around Hawaii to test for residual scan modulation effects and atmospheric correction biases. Chapter 6 provides a methodology for evaluating the atmospheric correction algorithm and atmospheric derived products using ground-based observations. Similarly, Chapter 7 presents match-up comparisons of coincident satellite and in situ data to determine the accuracy of the water-leaving radiances, chlorophyll a, and K(490) products.

  17. Calibration of well-type ionization chambers

    International Nuclear Information System (INIS)

    Alves, C.F.E.; Leite, S.P.; Pires, E.J.; Magalhaes, L.A.G.; David, M.G.; Almeida, C.E. de

    2015-01-01

    This paper presents the methodology developed by the Laboratorio de Ciencias Radiologicas and presently in use for determining of the calibration coefficient for well-type chambers used in the dosimetry of 192 Ir high dose rate sources. Uncertainty analysis involving the calibration procedure are discussed. (author)

  18. Construction of calibration curve for accountancy tank

    International Nuclear Information System (INIS)

    Kato, Takayuki; Goto, Yoshiki; Nidaira, Kazuo

    2009-01-01

    Tanks are equipped in a reprocessing plant for accounting solution of nuclear material. The careful measurement of volume in tanks is very important to implement rigorous accounting of nuclear material. The calibration curve relating the volume and level of solution needs to be constructed, where the level is determined by differential pressure of dip tubes. Several calibration curves are usually employed, but it's not explicitly decided how many segment are used, where to select segment, or what should be the degree of polynomial curve. These parameters, i.e., segment and degree of polynomial curve are mutually interrelated to give the better performance of calibration curve. Here we present the construction technique of giving optimum calibration curves and their characteristics. (author)

  19. Optimizing the accuracy of a helical diode array dosimeter: A comprehensive calibration methodology coupled with a novel virtual inclinometer

    Energy Technology Data Exchange (ETDEWEB)

    Kozelka, Jakub; Robinson, Joshua; Nelms, Benjamin; Zhang, Geoffrey; Savitskij, Dennis; Feygelman, Vladimir [Sun Nuclear Corp., Melbourne, Florida 32940 (United States); Department of Physics, University of South Florida, Tampa, Florida 33612 (United States); Canis Lupus LLC, Sauk County, Wisconsin 53561 (United States); Division of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612 (United States); Sun Nuclear Corp., Melbourne, Florida 32940 (United States); Division of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida 33612 (United States)

    2011-09-15

    Purpose: The goal of any dosimeter is to be as accurate as possible when measuring absolute dose to compare with calculated dose. This limits the uncertainties associated with the dosimeter itself and allows the task of dose QA to focus on detecting errors in the treatment planning (TPS) and/or delivery systems. This work introduces enhancements to the measurement accuracy of a 3D dosimeter comprised of a helical plane of diodes in a volumetric phantom. Methods: We describe the methods and derivations of new corrections that account for repetition rate dependence, intrinsic relative sensitivity per diode, field size dependence based on the dynamic field size determination, and positional correction. Required and described is an accurate ''virtual inclinometer'' algorithm. The system allows for calibrating the array directly against an ion chamber signal collected with high angular resolution. These enhancements are quantitatively validated using several strategies including ion chamber measurements taken using a ''blank'' plastic shell mimicking the actual phantom, and comparison to high resolution dose calculations for a variety of fields: static, simple arcs, and VMAT. A number of sophisticated treatment planning algorithms were benchmarked against ion chamber measurements for their ability to handle a large air cavity in the phantom. Results: Each calibration correction is quantified and presented vs its independent variable(s). The virtual inclinometer is validated by direct comparison to the gantry angle vs time data from machine log files. The effects of the calibration are quantified and improvements are seen in the dose agreement with the ion chamber reference measurements and with the TPS calculations. These improved agreements are a result of removing prior limitations and assumptions in the calibration methodology. Average gamma analysis passing rates for VMAT plans based on the AAPM TG-119 report are 98.4 and 93

  20. Calibration methodology application of kerma area product meters in situ: Preliminary results

    Science.gov (United States)

    Costa, N. A.; Potiens, M. P. A.

    2014-11-01

    The kerma-area product (KAP) is a useful quantity to establish the reference levels of conventional X-ray examinations. It can be obtained by measurements carried out with a KAP meter on a plane parallel transmission ionization chamber mounted on the X-ray system. A KAP meter can be calibrated in laboratory or in situ, where it is used. It is important to use one reference KAP meter in order to obtain reliable quantity of doses on the patient. The Patient Dose Calibrator (PDC) is a new equipment from Radcal that measures KAP. It was manufactured following the IEC 60580 recommendations, an international standard for KAP meters. This study had the aim to calibrate KAP meters using the PDC in situ. Previous studies and the quality control program of the PDC have shown that it has good function in characterization tests of dosimeters with ionization chamber and it also has low energy dependence. Three types of KAP meters were calibrated in four different diagnostic X-ray equipments. The voltages used in the two first calibrations were 50 kV, 70 kV, 100 kV and 120 kV. The other two used 50 kV, 70 kV and 90 kV. This was related to the equipments limitations. The field sizes used for the calibration were 10 cm, 20 cm and 30 cm. The calibrations were done in three different cities with the purpose to analyze the reproducibility of the PDC. The results gave the calibration coefficient for each KAP meter and showed that the PDC can be used as a reference instrument to calibrate clinical KAP meters.

  1. An overview of the dynamic calibration of piezoelectric pressure transducers

    Science.gov (United States)

    Theodoro, F. R. F.; Reis, M. L. C. C.; d’ Souto, C.

    2018-03-01

    Dynamic calibration is a research area that is still under development and is of great interest to aerospace and automotive industries. This study discusses some concepts regarding dynamic measurements of pressure quantities and presents an overview of dynamic calibration of pressure transducers. Studies conducted by the Institute of Aeronautics and Space focusing on research regarding piezoelectric pressure transducer calibration in shock tube are presented. We employed the Guide to the Expression of Uncertainty and a Monte Carlo Method in the methodology. The results show that both device and methodology employed are adequate to calibrate the piezoelectric sensor.

  2. Use of calibration methodology of gamma cameras for the workers surveillance using a thyroid simulator

    International Nuclear Information System (INIS)

    Alfaro, M.; Molina, G.; Vazquez, R.; Garcia, O.

    2010-09-01

    In Mexico there are a significant number of nuclear medicine centers in operation. For what the accidents risk related to the transport and manipulation of open sources used in nuclear medicine can exist. The National Institute of Nuclear Research (ININ) has as objective to establish a simple and feasible methodology for the workers surveillance related with the field of the nuclear medicine. This radiological surveillance can also be applied to the public in the event of a radiological accident. To achieve this it intends to use the available equipment s in the nuclear medicine centers, together with the neck-thyroid simulators elaborated by the ININ to calibrate the gamma cameras. The gamma cameras have among their component elements that conform spectrometric systems like the employees in the evaluation of the internal incorporation for direct measurements, reason why, besides their use for diagnostic for image, they can be calibrated with anthropomorphic simulators and also with punctual sources for the quantification of the radionuclides activity distributed homogeneously in the human body, or located in specific organs. Inside the project IAEA-ARCAL-RLA/9/049-LXXVIII -Procedures harmonization of internal dosimetry- where 9 countries intervened (Argentina, Brazil, Colombia, Cuba, Chile, Mexico, Peru, Uruguay and Spain). It was developed a protocol of cameras gamma calibration for the determination in vivo of radionuclides. The protocol is the base to establish and integrated network in Latin America to attend in response to emergencies, using nuclear medicine centers of public hospitals of the region. The objective is to achieve the appropriate radiological protection of the workers, essential for the sure and acceptable radiation use, the radioactive materials and the nuclear energy. (Author)

  3. Thermodynamic aspects of dynamical calibration of microbarometers used for IMS applications

    International Nuclear Information System (INIS)

    Starovojt, Yu.O.; Martysevich, P.N.; Kunakov, A.V.

    2006-01-01

    This paper describes the method of dynamical calibration of a microbarometer using the chamber with variable volume. The physics of this method is based on the thermodynamic process occurring inside the closed volume and on the corresponding relationships between pressure, volume and temperature changes caused by the movement of the piston attached to the inlet of the chamber. The method has been already used in several applications, however we consider essential to analyze its physical details. The paper gives the description of thermodynamic processes inside the calibration volume, the discussion of thermodynamics need and its effect on the calibration accuracy. (author)

  4. A methodology for calibration of hyperspectral and multispectral satellite data in coastal areas

    Science.gov (United States)

    Pennucci, Giuliana; Fargion, Giulietta; Alvarez, Alberto; Trees, Charles; Arnone, Robert

    2012-06-01

    The objective of this work is to determine the location(s) in any given oceanic area during different temporal periods where in situ sampling for Calibration/Validation (Cal/Val) provides the best capability to retrieve accurate radiometric and derived product data (lowest uncertainties). We present a method to merge satellite imagery with in situ measurements, to determine the best in situ sampling strategy suitable for satellite Cal/Val and to evaluate the present in situ locations through uncertainty indices. This analysis is required to determine if the present in situ sites are adequate for assessing uncertainty and where additional sites and ship programs should be located to improve Calibration/Validation (Cal/Val) procedures. Our methodology uses satellite acquisitions to build a covariance matrix encoding the spatial-temporal variability of the area of interest. The covariance matrix is used in a Bayesian framework to merge satellite and in situ data providing a product with lower uncertainty. The best in situ location for Cal/Val is then identified by using a design principle (A-optimum design) that looks for minimizing the estimated variance of the merged products. Satellite products investigated in this study include Ocean Color water leaving radiance, chlorophyll, and inherent and apparent optical properties (retrieved from MODIS and VIIRS). In situ measurements are obtained from systems operated on fixed deployment platforms (e.g., sites of the Ocean Color component of the AErosol RObotic NETwork- AERONET-OC), moorings (e.g, Marine Optical Buoy-MOBY), ships or autonomous vehicles (such as Autonomous Underwater Vehicles and/or Gliders).

  5. Airborne gamma-ray spectrometer and magnetometer survey. Volume I. Final report

    International Nuclear Information System (INIS)

    An airborne combined radiometric and magnetic survey was performed for the Department of Energy (DOE) over the area covered by the Mariposa, California and Nevada; Fresno, California; and Bakersfield, Caifornia 1:250,000 National Topographic Map Series (NTMS) 1 0 x 2 0 quadrangle maps. The survey was a part of DOE's National Aerial Radiometric Reconnaissance (ARR) program, which in turn is a part of the National Uranium Resource Evaluation (NURE) program. Data were collected by a helicopter equipped with a gamma-ray spectrometer having a large crystal volume, and a high sensitivity proton precession magnetometer. The radiometric system was calibrated at the Walker Field Calibration pads and the Lake Mead Dynamic Test range. Data quality was ensured throughout the survey by daily test flights and equipment checks. Radiometric data were corrected for live time, aircraft and equipment background, cosmic background, atmospheric radon, Compton scatter, and altitude dependence. The corrected data were statistically evaluated, plotted, and contoured to produce anomaly maps based on the radiometric response of individual geological units. These maps were interpreted and an anomaly interpretation map produced. Volume I contains a description of the systems used in the survey, a discussion of the calibration of the systems, the data processing procedures, the data display format, the interpretation rationale, and the interpretation methodology. A separate Volume II for each quadrangle contains the data displays and the interpretation results

  6. Overview of hypersonic CFD code calibration studies

    Science.gov (United States)

    Miller, Charles G.

    1987-01-01

    The topics are presented in viewgraph form and include the following: definitions of computational fluid dynamics (CFD) code validation; climate in hypersonics and LaRC when first 'designed' CFD code calibration studied was initiated; methodology from the experimentalist's perspective; hypersonic facilities; measurement techniques; and CFD code calibration studies.

  7. Investigation of a Fiber Optic Strain Sensing (FOSS) Distributed Load Calibration Methodology

    Data.gov (United States)

    National Aeronautics and Space Administration — FOSS is a relatively newer technology that needs to be explored for application to load calibration and loads monitoring efforts. Load calibration opportunities are...

  8. Automated Energy Calibration and Fitting of LaCl3(Ce y-Spectra Using Peak Likelihood and Tabu Search

    Directory of Open Access Journals (Sweden)

    Timothy P. McClanahan

    2008-10-01

    Full Text Available An automated method for ?-emission spectrum calibration and deconvolution is presented for spaceflight applications for a Cerium doped Lanthanum Chloride, (LaCl3(Ce ?-ray detector system. This detector will be coupled with a pulsed neutron generator (PNG to induce and enhance nuclide signal quality and rates, yielding large volumes of spectral information. Automated analytical methods are required to deconvolve and quantify nuclide signals from spectra; this will both reduce human interactions in spectrum analysis and facilitate feedback to automated robotic and operations planning. Initial system tests indicate significant energy calibration drifts (>6%, that which must be mitigated for spectrum analysis. A linear energy calibration model is presently considered, with gain and zero factors. Deconvolution methods incorporate a tabu search heuristic to formulate and optimize searches using memory structures. Iterative use of a peak likelihood methodology identifies global calibration minima and peak areas. The method is compared to manual methods of calibration and indicates superior performance using tabu methods. Performance of the Tabu enhanced calibration method is superior to similar unoptimized local search. The techniques are also applicable to other emission spectroscopy, eg. X-ray and neutron.

  9. SeaWiFS calibration and validation plan, volume 3

    International Nuclear Information System (INIS)

    Hooker, S.B.; Firestone, E.R.; Mcclain, C.R.; Esaias, W.E.; Barnes, W.; Guenther, B.; Endres, D.; Mitchell, B.G.; Barnes, R.

    1992-09-01

    The Sea-viewing Wide Field-of-view Sensor (SeaWiFS) will be the first ocean-color satellite since the Nimbus-7 Coastal Zone Color Scanner (CZCS), which ceased operation in 1986. Unlike the CZCS, which was designed as a proof-of-concept experiment, SeaWiFS will provide routine global coverage every 2 days and is designed to provide estimates of photosynthetic concentrations of sufficient accuracy for use in quantitative studies of the ocean's primary productivity and biogeochemistry. A review of the CZCS mission is included that describes that data set's limitations and provides justification for a comprehensive SeaWiFS calibration and validation program. To accomplish the SeaWiFS scientific objectives, the sensor's calibration must be constantly monitored, and robust atmospheric corrections and bio-optical algorithms must be developed. The plan incorporates a multi-faceted approach to sensor calibration using a combination of vicarious (based on in situ observations) and onboard calibration techniques. Because of budget constraints and the limited availability of ship resources, the development of the operational algorithms (atmospheric and bio-optical) will rely heavily on collaborations with the Earth Observing System (EOS), the Moderate Resolution Imaging Spectrometer (MODIS) oceans team, and projects sponsored by other agencies, e.g., the U.S. Navy and the National Science Foundation (NSF). Other elements of the plan include the routine quality control of input ancillary data (e.g., surface wind, surface pressure, ozone concentration, etc.) used in the processing and verification of the level-0 (raw) data to level-1 (calibrated radiances), level-2 (derived products), and level-3 (gridded and averaged derived data) products

  10. Genetic algorithm for building envelope calibration

    International Nuclear Information System (INIS)

    Ramos Ruiz, Germán; Fernández Bandera, Carlos; Gómez-Acebo Temes, Tomás; Sánchez-Ostiz Gutierrez, Ana

    2016-01-01

    Highlights: • Calibration methodology using Multi-Objective Genetic Algorithm (NSGA-II). • Uncertainty analysis formulas implemented directly in EnergyPlus. • The methodology captures the heat dynamic of the building with a high level of accuracy. • Reduction in the number of parameters involved due to sensitivity analysis. • Cost-effective methodology using temperature sensors only. - Abstract: Buildings today represent 40% of world primary energy consumption and 24% of greenhouse gas emissions. In our society there is growing interest in knowing precisely when and how energy consumption occurs. This means that consumption measurement and verification plans are well-advanced. International agencies such as Efficiency Valuation Organization (EVO) and International Performance Measurement and Verification Protocol (IPMVP) have developed methodologies to quantify savings. This paper presents a methodology to accurately perform automated envelope calibration under option D (calibrated simulation) of IPMVP – vol. 1. This is frequently ignored because of its complexity, despite being more flexible and accurate in assessing the energy performance of a building. A detailed baseline energy model is used, and by means of a metaheuristic technique achieves a highly reliable and accurate Building Energy Simulation (BES) model suitable for detailed analysis of saving strategies. In order to find this BES model a Genetic Algorithm (NSGA-II) is used, together with a highly efficient engine to stimulate the objective, thus permitting rapid achievement of the goal. The result is a BES model that broadly captures the heat dynamic behaviour of the building. The model amply fulfils the parameters demanded by ASHRAE and EVO under option D.

  11. NLT and extrapolated DLT:3-D cinematography alternatives for enlarging the volume of calibration.

    Science.gov (United States)

    Hinrichs, R N; McLean, S P

    1995-10-01

    This study investigated the accuracy of the direct linear transformation (DLT) and non-linear transformation (NLT) methods of 3-D cinematography/videography. A comparison of standard DLT, extrapolated DLT, and NLT calibrations showed the standard (non-extrapolated) DLT to be the most accurate, especially when a large number of control points (40-60) were used. The NLT was more accurate than the extrapolated DLT when the level of extrapolation exceeded 100%. The results indicated that when possible one should use the DLT with a control object, sufficiently large as to encompass the entire activity being studied. However, in situations where the activity volume exceeds the size of one's DLT control object, the NLT method should be considered.

  12. A single model procedure for estimating tank calibration equations

    International Nuclear Information System (INIS)

    Liebetrau, A.M.

    1997-10-01

    A fundamental component of any accountability system for nuclear materials is a tank calibration equation that relates the height of liquid in a tank to its volume. Tank volume calibration equations are typically determined from pairs of height and volume measurements taken in a series of calibration runs. After raw calibration data are standardized to a fixed set of reference conditions, the calibration equation is typically fit by dividing the data into several segments--corresponding to regions in the tank--and independently fitting the data for each segment. The estimates obtained for individual segments must then be combined to obtain an estimate of the entire calibration function. This process is tedious and time-consuming. Moreover, uncertainty estimates may be misleading because it is difficult to properly model run-to-run variability and between-segment correlation. In this paper, the authors describe a model whose parameters can be estimated simultaneously for all segments of the calibration data, thereby eliminating the need for segment-by-segment estimation. The essence of the proposed model is to define a suitable polynomial to fit to each segment and then extend its definition to the domain of the entire calibration function, so that it (the entire calibration function) can be expressed as the sum of these extended polynomials. The model provides defensible estimates of between-run variability and yields a proper treatment of between-segment correlations. A portable software package, called TANCS, has been developed to facilitate the acquisition, standardization, and analysis of tank calibration data. The TANCS package was used for the calculations in an example presented to illustrate the unified modeling approach described in this paper. With TANCS, a trial calibration function can be estimated and evaluated in a matter of minutes

  13. The Fundamentals of the Air Sampler Calibration-Verification Process

    International Nuclear Information System (INIS)

    Gavila, F.M.

    2011-01-01

    The calibration of an air sampling instrument using a reference air flow calibrator requires attention to scientific detail in order to establish that the instrument's reported values are correctly stated and valid under the actual operating conditions of the air sampling instrument. The primary objective of an air flow calibration-verification is to ensure that the device under test (DUT) is within the manufacturer's stated accuracy range of temperature, pressure and humidity conditions under which the instrument was designed to operate. The DUT output values are compared to those obtained from a reference instrument (REF) measuring the sample physical parameter that the DUT is measuring. An accurate comparison of air flow rates or air volumes requires that the comparison of the DUT and REF values be made under the same temperature and pressure conditions. It is absolutely necessary that the REF be more accurate than the DUT; otherwise, it can not be considered a reference instrument. The REF should be at least twice as accurate and, if possible, it should be four times as accurate as the DUT. Upon confirmation that the DUT meets the manufacturer's accuracy criteria, the technician must place a calibration sticker or label indicating the date of calibration, the expiration date of the calibration and an authorized signature. If it is a limited-use instrument, the label should state the limited-use operating range. The serial number and model number of the instrument should also be shown on the calibration sticker. A specific calibration file for each instrument by serial number should be kept in the calibration laboratory file records. Instruments that display gas flow or gas volume values corrected to a reference temperature and pressure are very desirable. The ideal situation is when both the DUT and the REF output flow rate or volume values are at the same conditions of T and P. The calibration-verification is, then, a simple process. The credibility of an air

  14. The Self-Concept. Volume 1, A Review of Methodological Considerations and Measuring Instruments. Revised Edition.

    Science.gov (United States)

    Wylie, Ruth C.

    This volume of the revised edition describes and evaluates measurement methods, research designs, and procedures which have been or might appropriately be used in self-concept research. Working from the perspective that self-concept or phenomenal personality theories can be scientifically investigated, methodological flaws and questionable…

  15. Cumulative error models for the tank calibration problem

    International Nuclear Information System (INIS)

    Goldman, A.; Anderson, L.G.; Weber, J.

    1983-01-01

    The purpose of a tank calibration equation is to obtain an estimate of the liquid volume that corresponds to a liquid level measurement. Calibration experimental errors occur in both liquid level and liquid volume measurements. If one of the errors is relatively small, the calibration equation can be determined from wellknown regression and calibration methods. If both variables are assumed to be in error, then for linear cases a prototype model should be considered. Many investigators are not familiar with this model or do not have computing facilities capable of obtaining numerical solutions. This paper discusses and compares three linear models that approximate the prototype model and have the advantage of much simpler computations. Comparisons among the four models and recommendations of suitability are made from simulations and from analyses of six sets of experimental data

  16. Quality of the neutron probe calibration curve

    International Nuclear Information System (INIS)

    Libardi, Paulo Leonel; Moraes, Sergio Oliveira

    1997-01-01

    An experiment of neutron probe calibration has been performed, involving various volume size samples and collected at various distances from the access tubes. The experiment aimed to give some answers to questions such as suitable sample physical volume, always use of the same volume and sample distance from the neutron probe access tube

  17. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-07-01

    Accurate solar radiation data sets are critical to reducing the expenses associated with mitigating performance risk for solar energy conversion systems, and they help utility planners and grid system operators understand the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of calibration methodologies and the resulting calibration responsivities provided by radiometric calibration service providers such as the National Renewable Energy Laboratory (NREL) and manufacturers of radiometers. Some of these radiometers are calibrated indoors, and some are calibrated outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The reference radiometer calibrations are traceable to the World Radiometric Reference. These different methods of calibration demonstrated 1% to 2% differences in solar irradiance measurement. Analyzing these values will ultimately assist in determining the uncertainties of the radiometer data and will assist in developing consensus on a standard for calibration.

  18. Development of a calibration system for surface contamination monitors

    International Nuclear Information System (INIS)

    Marechal, M.H.H.; Barbosa, M.P.

    1992-01-01

    A calibration system for surface contamination monitors is developed, aiming supply the existence demand of these instruments. A experimental arrangement and a methodology are described. The advantages of use this system for calibration routine optimization are also discussed. (C.G.C.)

  19. Infrared stereo calibration for unmanned ground vehicle navigation

    Science.gov (United States)

    Harguess, Josh; Strange, Shawn

    2014-06-01

    The problem of calibrating two color cameras as a stereo pair has been heavily researched and many off-the-shelf software packages, such as Robot Operating System and OpenCV, include calibration routines that work in most cases. However, the problem of calibrating two infrared (IR) cameras for the purposes of sensor fusion and point could generation is relatively new and many challenges exist. We present a comparison of color camera and IR camera stereo calibration using data from an unmanned ground vehicle. There are two main challenges in IR stereo calibration; the calibration board (material, design, etc.) and the accuracy of calibration pattern detection. We present our analysis of these challenges along with our IR stereo calibration methodology. Finally, we present our results both visually and analytically with computed reprojection errors.

  20. Application of methodology for calibration of instruments utilized in dosimetry of high energy beams, for radiodiagnosis

    International Nuclear Information System (INIS)

    Potiens, Maria P.A.; Caldas, Linda V.E.

    2000-01-01

    The radiation qualities recommended by the IEC 1267 standard for the calibration of instruments used in diagnostic radiology measurements were established using a neo-diagnomax X-ray system (125 kV). The RQR radiation qualities are recommended to test ionization chambers used in non attenuated beams, and the RQA radiation qualities in attenuated beams (behind a phantom). To apply the methodology, 6 ionization chambers commonly used in diagnostic radiology were tested. The higher energy dependence (17%) was obtained for an ionization chamber recommended for mammography beams, that is not the case of the X radiation system used in this work. The other ionization chambers presented good performance in terms of energy (maximum of 5%), therefore within the limits of the international recommendations for this kind of instrument. (author)

  1. Calibration of the SNO+ experiment

    Science.gov (United States)

    Maneira, J.; Falk, E.; Leming, E.; Peeters, S.; SNO+ Collaboration.

    2017-09-01

    The main goal of the SNO+ experiment is to perform a low-background and high-isotope-mass search for neutrinoless double-beta decay, employing 780 tonnes of liquid scintillator loaded with tellurium, in its initial phase at 0.5% by mass for a total mass of 1330 kg of 130Te. The SNO+ physics program includes also measurements of geo- and reactor neutrinos, supernova and solar neutrinos. Calibrations are an essential component of the SNO+ data-taking and analysis plan. The achievement of the physics goals requires both an extensive and regular calibration. This serves several goals: the measurement of several detector parameters, the validation of the simulation model and the constraint of systematic uncertainties on the reconstruction and particle identification algorithms. SNO+ faces stringent radiopurity requirements which, in turn, largely determine the materials selection, sealing and overall design of both the sources and deployment systems. In fact, to avoid frequent access to the inner volume of the detector, several permanent optical calibration systems have been developed and installed outside that volume. At the same time, the calibration source internal deployment system was re-designed as a fully sealed system, with more stringent material selection, but following the same working principle as the system used in SNO. This poster described the overall SNO+ calibration strategy, discussed the several new and innovative sources, both optical and radioactive, and covered the developments on source deployment systems.

  2. Modelling of thermal hydraulics in a KAROLINA calorimeter for its calibration methodology validation

    Directory of Open Access Journals (Sweden)

    Luks Aleksandra

    2016-12-01

    Full Text Available Results of numerical calculations of heat exchange in a nuclear heating detector for nuclear reactors are presented in this paper. The gamma radiation is generated in nuclear reactor during fission and radiative capture reactions as well as radioactive decay of its products. A single-cell calorimeter has been designed for application in the MARIA research reactor in the National Centre for Nuclear Research (NCBJ in Świerk near Warsaw, Poland, and can also be used in the Jules Horowitz Reactor (JHR, which is under construction in the research centre in Cadarache, France. It consists of a cylindrical sample, which is surrounded by a gas layer, contained in a cylindrical housing. Additional calculations had to be performed before its insertion into the reactor. Within this analysis, modern computational fluid dynamics (CFD methods have been used for assessing important parameters, for example, mean surface temperature, mean volume temperature, and maximum sample (calorimeter core temperature. Results of an experiment performed at a dedicated out-of-pile calibration bench and results of numerical modelling validation are also included in this paper.

  3. BES online calibration system

    International Nuclear Information System (INIS)

    Zhang Bingyun; Li Xiaonan; Zhu Kejun; Zhang Jiawen; Gong Mingyu

    2003-01-01

    We constructed BES (Beijing Spectrometer) online calibration system to ensure the coherence of readout electronic channels due to huge data volume in high energy physics experiment. This paper describes the structure of hardware and software, and its characteristic and function

  4. GO methodology. Volume 1. Overview manual

    International Nuclear Information System (INIS)

    1983-06-01

    The GO methodology is a success-oriented probabilistic system performance analysis technique. The methodology can be used to quantify system reliability and availability, identify and rank critical components and the contributors to system failure, construct event trees, and perform statistical uncertainty analysis. Additional capabilities of the method currently under development will enhance its use in evaluating the effects of external events and common cause failures on system performance. This Overview Manual provides a description of the GO Methodology, how it can be used, and benefits of using it in the analysis of complex systems

  5. Calibration of modified parallel-plate rheometer through calibrated oil and lattice Boltzmann simulation

    DEFF Research Database (Denmark)

    Ferraris, Chiara F; Geiker, Mette Rica; Martys, Nicos S

    2007-01-01

    inapplicable here. This paper presents the analysis of a modified parallel plate rheometer for measuring cement mortar and propose a methodology for calibration using standard oils and numerical simulation of the flow. A lattice Boltzmann method was used to simulate the flow in the modified rheometer, thus...

  6. A rational methodology for the study of foundations for marine structures; Una metdologia racional para el estudio de cimentaciones de estructuras marinas

    Energy Technology Data Exchange (ETDEWEB)

    Mira Mc Willams, P.; Fernandez-Merodo, J. A.; Pastor Perez, M.; Monte Saez, J. L.; Martinez Santamaria, J. M.; Cuellar Mirasol, V.; Martin Baanante, M. E.; Rodriguez Sanchez-Arevalo, I; Lopez Maldonando, J. D.; Tomas Sampedro, A.

    2011-07-01

    A methodology for the study of marine foundations is presented. The response in displacements, stresses and pore water pressures in obtained from a finite element coupled formulation. Loads due to wave action of the foundation are obtained from a volume of fluid type fluid-structure interaction numerical model. Additionally, the methodology includes a Generalized Plasticity based constitutive model for granular materials capable of representing liquefaction fenomena of sands subjected to cyclic loading, such as those frequently appearing in the problems studied. Calibration of this model requires a series of laboratory tests detailed herein. This methodology is applied to the study of the response of a caisson breakwater foundation. (Author) 10 refs.

  7. Development of a calibration methodology and tests of kerma area product meters; Desenvolvimento de uma metodologia de calibracao e testes de medidores de produto Kerma-Area

    Energy Technology Data Exchange (ETDEWEB)

    Costa, Nathalia Almeida

    2013-07-01

    The quantity kerma area product (PKA) is important to establish reference levels in diagnostic radiology exams. This quantity can be obtained using a PKA meter. The use of such meters is essential to evaluate the radiation dose in radiological procedures and is a good indicator to make sure that the dose limit to the patient's skin doesn't exceed. Sometimes, these meters come fixed to X radiation equipment, which makes its calibration difficult. In this work, it was developed a methodology for calibration of PKA meters. The instrument used for this purpose was the Patient Dose Calibrator (PDC). It was developed to be used as a reference to check the calibration of PKA and air kerma meters that are used for dosimetry in patients and to verify the consistency and behavior of systems of automatic exposure control. Because it is a new equipment, which, in Brazil, is not yet used as reference equipment for calibration, it was also performed the quality control of this equipment with characterization tests, the calibration and an evaluation of the energy dependence. After the tests, it was proved that the PDC can be used as a reference instrument and that the calibration must be performed in situ, so that the characteristics of each X-ray equipment, where the PKA meters are used, are considered. The calibration was then performed with portable PKA meters and in an interventional radiology equipment that has a PKA meter fixed. The results were good and it was proved the need for calibration of these meters and the importance of in situ calibration with a reference meter. (author)

  8. Airborne gamma-ray spectrometer and magnetometer survey: Lund quadrangle, Ely quadrangle, Nevada. Volume I. Final report

    International Nuclear Information System (INIS)

    1980-01-01

    An airborne combined radiometric and magnetic survey was performed for the Department of Energy (DOE) over the area covered by the Ely and Lund 1:250,000 National Topographic Map Series (NTMS quadrangle maps). The survey was part of DOE's National Uranium Resource Evaluation (NURE) Aerial Radiometric Reconnaissance program. Data were collected by a helicopter equipped with a gamma-ray spectrometer having a large crystal volume, and a high sensitivity proton precession magnetometer. The radiometric system was calibrated at the Walker Field Calibration pads and the Lake Mead Dynamic Test range. Data quality was ensured during the survey by daily test flights and equipment checks. Radiometric data were corrected for live time, aircraft and equipment background, cosmic background, atmospheric radon, Compton scatter, and altitude dependence. The corrected data were statistically evaluated, plotted, and contoured to produce anomaly maps based on the radiometric response of individual geological units. The maps were interpreted and an anomaly interpretation map produced. Volume I contains a description of the systems used in the survey, a discussion of the calibration of the systems, the data processing procedures, the data display format, the interpretation rationale, and interpretation methodology. Volume II contains the data displays for a quadrangle and the interpretation results

  9. On the development of a new methodology in sub-surface parameterisation on the calibration of groundwater models

    Science.gov (United States)

    Klaas, D. K. S. Y.; Imteaz, M. A.; Sudiayem, I.; Klaas, E. M. E.; Klaas, E. C. M.

    2017-10-01

    In groundwater modelling, robust parameterisation of sub-surface parameters is crucial towards obtaining an agreeable model performance. Pilot point is an alternative in parameterisation step to correctly configure the distribution of parameters into a model. However, the methodology given by the current studies are considered less practical to be applied on real catchment conditions. In this study, a practical approach of using geometric features of pilot point and distribution of hydraulic gradient over the catchment area is proposed to efficiently configure pilot point distribution in the calibration step of a groundwater model. A development of new pilot point distribution, Head Zonation-based (HZB) technique, which is based on the hydraulic gradient distribution of groundwater flow, is presented. Seven models of seven zone ratios (1, 5, 10, 15, 20, 25 and 30) using HZB technique were constructed on an eogenetic karst catchment in Rote Island, Indonesia and their performances were assessed. This study also concludes some insights into the trade-off between restricting and maximising the number of pilot points and offers a new methodology for selecting pilot point properties and distribution method in the development of a physically-based groundwater model.

  10. Comparison of Two Methodologies for Calibrating Satellite Instruments in the Visible and Near-Infrared

    Science.gov (United States)

    Barnes, Robert A.; Brown, Steven W.; Lykke, Keith R.; Guenther, Bruce; Butler, James J.; Schwarting, Thomas; Turpie, Kevin; Moyer, David; DeLuccia, Frank; Moeller, Christopher

    2015-01-01

    Traditionally, satellite instruments that measure Earth-reflected solar radiation in the visible and near infrared wavelength regions have been calibrated for radiance responsivity in a two-step method. In the first step, the relative spectral response (RSR) of the instrument is determined using a nearly monochromatic light source such as a lamp-illuminated monochromator. These sources do not typically fill the field-of-view of the instrument nor act as calibrated sources of light. Consequently, they only provide a relative (not absolute) spectral response for the instrument. In the second step, the instrument views a calibrated source of broadband light, such as a lamp-illuminated integrating sphere. The RSR and the sphere absolute spectral radiance are combined to determine the absolute spectral radiance responsivity (ASR) of the instrument. More recently, a full-aperture absolute calibration approach using widely tunable monochromatic lasers has been developed. Using these sources, the ASR of an instrument can be determined in a single step on a wavelength-by-wavelength basis. From these monochromatic ASRs, the responses of the instrument bands to broadband radiance sources can be calculated directly, eliminating the need for calibrated broadband light sources such as lamp-illuminated integrating spheres. In this work, the traditional broadband source-based calibration of the Suomi National Preparatory Project (SNPP) Visible Infrared Imaging Radiometer Suite (VIIRS) sensor is compared with the laser-based calibration of the sensor. Finally, the impact of the new full-aperture laser-based calibration approach on the on-orbit performance of the sensor is considered.

  11. Calibration and evaluation of neutron moisturemeter

    International Nuclear Information System (INIS)

    Tang Zhangxiong; Hu Jiangchao; Sun Laiyan; Wang Huaihui; Wu Weixue

    1992-02-01

    Factors influencing the calibration curve of neutron moisture meter, such as soil type, texture, volume weight and depth, were studied. When the soil bulk density water content is between 15% to 45%, the calibration curve is approximately a straight line, and the intercept and slope are only influenced by the above factors. The growing plants also influence the calibration curve slightly. The measuring error for top soil (< 20 cm) is larger. The relative error between neutron method and weighing method is about 8%. The neutron method has many advantages such as non-interfering, simple, fast and non-time-delay

  12. Calibrating the neutron moisture meter: Precision and economy

    International Nuclear Information System (INIS)

    Akhter, J.; Waheed, R.A.; Hignett, C.T.; Greacen, E.L.

    2000-01-01

    Established laboratory and field calibration procedures for the neutron moisture meter are demonstrated on a uniform soil and alternative, low cost procedures on a duplex, less uniform soil. The effect of field variability on the calibration methodology is discussed with the aim of optimising calibration reliability at minimal cost. The difference between calibration for a soil material, or for a field (a range of soil materials) is considered. In particular, calibration for the estimation of water content change is shown to be a different problem from calibration for the estimation of water content in a variable field. Techniques aimed at detecting field variability problems during calibration are suggested, and methods for optimising the results for the intended use of the instrument are outlined. Pairing of calibration tubes, alternative methods of analysis of calibration data, and use of other information from the field to measure its variability, can improve the precision of calibration procedures to the point where minimal calibration effort, with careful analysis, can provide reliable results. (author)

  13. Volume measurement study for large scale input accountancy tank

    International Nuclear Information System (INIS)

    Uchikoshi, Seiji; Watanabe, Yuichi; Tsujino, Takeshi

    1999-01-01

    Large Scale Tank Calibration (LASTAC) facility, including an experimental tank which has the same volume and structure as the input accountancy tank of Rokkasho Reprocessing Plant (RRP) was constructed in Nuclear Material Control Center of Japan. Demonstration experiments have been carried out to evaluate a precision of solution volume measurement and to establish the procedure of highly accurate pressure measurement for a large scale tank with dip-tube bubbler probe system to be applied to the input accountancy tank of RRP. Solution volume in a tank is determined from substitution the solution level for the calibration function obtained in advance, which express a relation between the solution level and its volume in the tank. Therefore, precise solution volume measurement needs a precise calibration function that is determined carefully. The LASTAC calibration experiments using pure water showed good result in reproducibility. (J.P.N.)

  14. Natural Computing in Computational Finance Volume 4

    CERN Document Server

    O’Neill, Michael; Maringer, Dietmar

    2012-01-01

    This book follows on from Natural Computing in Computational Finance  Volumes I, II and III.   As in the previous volumes of this series, the  book consists of a series of  chapters each of  which was selected following a rigorous, peer-reviewed, selection process.  The chapters illustrate the application of a range of cutting-edge natural  computing and agent-based methodologies in computational finance and economics.  The applications explored include  option model calibration, financial trend reversal detection, enhanced indexation, algorithmic trading,  corporate payout determination and agent-based modeling of liquidity costs, and trade strategy adaptation.  While describing cutting edge applications, the chapters are  written so that they are accessible to a wide audience. Hence, they should be of interest  to academics, students and practitioners in the fields of computational finance and  economics.  

  15. Methodology for the development and calibration of the SCI-QOL item banks.

    Science.gov (United States)

    Tulsky, David S; Kisala, Pamela A; Victorson, David; Choi, Seung W; Gershon, Richard; Heinemann, Allen W; Cella, David

    2015-05-01

    To develop a comprehensive, psychometrically sound, and conceptually grounded patient reported outcomes (PRO) measurement system for individuals with spinal cord injury (SCI). Individual interviews (n=44) and focus groups (n=65 individuals with SCI and n=42 SCI clinicians) were used to select key domains for inclusion and to develop PRO items. Verbatim items from other cutting-edge measurement systems (i.e. PROMIS, Neuro-QOL) were included to facilitate linkage and cross-population comparison. Items were field tested in a large sample of individuals with traumatic SCI (n=877). Dimensionality was assessed with confirmatory factor analysis. Local item dependence and differential item functioning were assessed, and items were calibrated using the item response theory (IRT) graded response model. Finally, computer adaptive tests (CATs) and short forms were administered in a new sample (n=245) to assess test-retest reliability and stability. A calibration sample of 877 individuals with traumatic SCI across five SCI Model Systems sites and one Department of Veterans Affairs medical center completed SCI-QOL items in interview format. We developed 14 unidimensional calibrated item banks and 3 calibrated scales across physical, emotional, and social health domains. When combined with the five Spinal Cord Injury--Functional Index physical function banks, the final SCI-QOL system consists of 22 IRT-calibrated item banks/scales. Item banks may be administered as CATs or short forms. Scales may be administered in a fixed-length format only. The SCI-QOL measurement system provides SCI researchers and clinicians with a comprehensive, relevant and psychometrically robust system for measurement of physical-medical, physical-functional, emotional, and social outcomes. All SCI-QOL instruments are freely available on Assessment CenterSM.

  16. Maximum respiratory pressure measuring system : calibration and evaluation of uncertainty

    NARCIS (Netherlands)

    Ferreira, J.L.; Pereira, N.C.; Oliveira Júnior, M.; Vasconcelos, F.H.; Parreira, V.F.; Tierra-Criollo, C.J.

    2010-01-01

    The objective of this paper is to present a methodology for the evaluation of uncertainties in the measurements results obtained during the calibration of a digital manovacuometer prototype (DM) with a load cell sensor pressure device incorporated. Calibration curves were obtained for both pressure

  17. Marine X-band Weather Radar Data Calibration

    DEFF Research Database (Denmark)

    Thorndahl, Søren Liedtke; Rasmussen, Michael R.

    2012-01-01

    estimates. This paper presents some of the challenges in small marine X-band radar calibration by comparing three calibration procedures for assessing the relationship between radar and rain gauge data. Validation shows similar results for precipitation volumes but more diverse results on peak rain......Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis, and real time control purposes. In these contexts, it is allimportant that the radar data is well calibrated and adjusted in order to obtain valid quantitative precipitation...

  18. Optimization of procedure for calibration with radiometer/photometer

    International Nuclear Information System (INIS)

    Detilly, Isabelle

    2009-01-01

    A test procedure for the radiometer/photometer calibrations mark International Light at the Laboratorio de Fotometria y Tecnologia Laser (LAFTA) de la Escuela de Ingenieria Electrica de la Universidad de Costa Rica is established. Two photometric banks are used as experimental set and two calibrations were performed of the International Light. A basic procedure established in the laboratory, is used for calibration from measurements of illuminance and luminous intensity. Some dependent variations of photometric banks used in the calibration process, the programming of the radiometer/photometer and the applied methodology showed the results. The procedure for calibration with radiometer/photometer can be improved by optimizing the programming process of the measurement instrument and possible errors can be minimized by using the recommended procedure. (author) [es

  19. Selection methodology for LWR safety programs and proposals. Volume 2. Methodology application

    International Nuclear Information System (INIS)

    Ritzman, R.L.; Husseiny, A.A.

    1980-08-01

    The results of work done to update and apply a methodology for selecting (prioritizing) LWR safety technology R and D programs are described. The methodology is based on multiattribute utility (MAU) theory. Application of the methodology to rank-order a group of specific R and D programs included development of a complete set of attribute utility functions, specification of individual attribute scaling constants, and refinement and use of an interactive computer program (MAUP) to process decision-maker inputs and generate overall (multiattribute) program utility values. The output results from several decision-makers are examined for consistency and conclusions and recommendations regarding general use of the methodology are presented. 3 figures, 18 tables

  20. Methodology to Forecast Volume and Cost of Cancer Drugs in Low- and Middle-Income Countries

    Directory of Open Access Journals (Sweden)

    Yehoda M. Martei

    2018-02-01

    Full Text Available Purpose: In low- and middle-income countries (LMICs, frequent outages of the stock of cancer drugs undermine cancer care delivery and are potentially fatal for patients with cancer. The aim of this study is to describe a methodologic approach to forecast chemotherapy volume and estimate cost that can be readily updated and applied in most LMICs. Methods: Prerequisite data for forecasting are population-based incidence data and cost estimates per unit of drug to be ordered. We used the supplementary guidelines from the WHO list of essential medicines for cancer to predict treatment plans and ordering patterns. We used de-identified aggregate data from the Botswana National Cancer Registry to estimate incident cases. The WHO Management Sciences for Health International Price Indicator was used to estimate unit costs per drug. Results: Chemotherapy volume required for incident cancer cases was estimated as the product of the standardized dose required to complete a full treatment regimen per patient, with a given cancer diagnosis and stage, multiplied by the total number of incident cancer cases with the respective diagnosis. The estimated chemotherapy costs to treat the 10 most common cancers in the public health care sector of Botswana is approximately 2.3 million US dollars. An estimated 66% of the budget is allocated to costs of rituximab and trastuzumab alone, which are used by approximately 10% of the cancer population. Conclusion: This method provides a reproducible approach to forecast chemotherapy volume and cost in LMICs. The chemotherapy volume and cost outputs of this methodology provide key stakeholders with valuable information that can guide budget estimation, resource allocation, and drug-price negotiations for cancer treatment. Ultimately, this will minimize drug shortages or outages and reduce potential loss of lives that result from an erratic drug supply.

  1. Vessel calibration for accurate material accountancy at RRP

    International Nuclear Information System (INIS)

    Yanagisawa, Yuu; Ono, Sawako; Iwamoto, Tomonori

    2004-01-01

    RRP has a 800t·Upr capacity a year to re-process, where would be handled a large amount of nuclear materials as solution. A large scale plant like RRP will require accurate materials accountancy system, so that the vessel calibration with high-precision is very important as initial vessel calibration before operation. In order to obtain the calibration curve, it is needed well-known each the increment volume related with liquid height. Then we performed at least 2 or 3 times run with water for vessel calibration and careful evaluation for the calibration data should be needed. We performed vessel calibration overall 210 vessels, and the calibration of 81 vessels including IAT and OAT were held under presence of JSGO and IAEA inspectors taking into account importance on the material accountancy. This paper describes outline of the initial vessel calibration and calibration results based on back pressure measurement with dip tubes. (author)

  2. In situ calibration of TFTR neutron detectors

    International Nuclear Information System (INIS)

    Hendel, H.W.; Palladino, R.W.; Barnes, C.W.; Diesso, M.; Felt, J.S.; Jassby, D.L.; Johnson, L.C.; Ku, L.; Liu, Q.P.; Motley, R.W.; Murphy, H.B.; Murphy, J.; Nieschmidt, E.B.; Roberts, J.A.; Saito, T.; Strachan, J.D.; Waszazak, R.J.; Young, K.M.

    1990-01-01

    We report results of the TFTR fission detector calibration performed in December 1988. A NBS-traceable, remotely controlled 252 Cf neutron source was moved toroidally through the TFTR vacuum vessel. Detection efficiencies for two 235 U detectors were measured for 930 locations of the neutron point source in toroidal scans at 16 different major radii and vertical heights. These scans effectively simulated the volume-distributed plasma neutron source and the volume-integrated detection efficiency was found to be insensitive to plasma position. The Campbell mode is useful due to its large overlap with the count rate mode and large dynamic range. The resulting absolute plasma neutron source calibration has an uncertainty of ±13%

  3. Adaptive calibration of (u,v)‐wind ensemble forecasts

    DEFF Research Database (Denmark)

    Pinson, Pierre

    2012-01-01

    of sufficient reliability. The original framework introduced here allows for an adaptive bivariate calibration of these ensemble forecasts. The originality of this methodology lies in the fact that calibrated ensembles still consist of a set of (space–time) trajectories, after translation and dilation...... of translation and dilation factors are discussed. Copyright © 2012 Royal Meteorological Society...

  4. Calibration requirements and methodology for remote sensors viewing the ocean in the visible

    Science.gov (United States)

    Gordon, Howard R.

    1987-01-01

    The calibration requirements for ocean-viewing sensors are outlined, and the present methods of effecting such calibration are described in detail. For future instruments it is suggested that provision be made for the sensor to view solar irradiance in diffuse reflection and that the moon be used as a source of diffuse light for monitoring the sensor stability.

  5. Calibration Uncertainties in the Droplet Measurement Technologies Cloud Condensation Nuclei Counter

    Science.gov (United States)

    Hibert, Kurt James

    Cloud condensation nuclei (CCN) serve as the nucleation sites for the condensation of water vapor in Earth's atmosphere and are important for their effect on climate and weather. The influence of CCN on cloud radiative properties (aerosol indirect effect) is the most uncertain of quantified radiative forcing changes that have occurred since pre-industrial times. CCN influence the weather because intrinsic and extrinsic aerosol properties affect cloud formation and precipitation development. To quantify these effects, it is necessary to accurately measure CCN, which requires accurate calibrations using a consistent methodology. Furthermore, the calibration uncertainties are required to compare measurements from different field projects. CCN uncertainties also aid the integration of CCN measurements with atmospheric models. The commercially available Droplet Measurement Technologies (DMT) CCN Counter is used by many research groups, so it is important to quantify its calibration uncertainty. Uncertainties in the calibration of the DMT CCN counter exist in the flow rate and supersaturation values. The concentration depends on the accuracy of the flow rate calibration, which does not have a large (4.3 %) uncertainty. The supersaturation depends on chamber pressure, temperature, and flow rate. The supersaturation calibration is a complex process since the chamber's supersaturation must be inferred from a temperature difference measurement. Additionally, calibration errors can result from the Kohler theory assumptions, fitting methods utilized, the influence of multiply-charged particles, and calibration points used. In order to determine the calibration uncertainties and the pressure dependence of the supersaturation calibration, three calibrations are done at each pressure level: 700, 840, and 980 hPa. Typically 700 hPa is the pressure used for aircraft measurements in the boundary layer, 840 hPa is the calibration pressure at DMT in Boulder, CO, and 980 hPa is the

  6. Effects of gypsum and bulk density on neutron probe calibration curves

    International Nuclear Information System (INIS)

    Arslan, Awadis; Razzouk, A.K.

    1993-10-01

    The effects of gypsum and bulk density on the neutron probe calibration curve were studied in the laboratory and in the field. The effect of bulk density was negligible for the soil studied in the laboratory, while it was significant for the field calibration. An increase in the slope of moisture content on a volume basis vs. count ratio with increasing gypsum content at the soil was observed in the laboratory calibration. A simple method for correction of the calibration curve for gypsum content was adopted to obtain a specific curve for each layer. The adapted method requires the gypsum fraction to be estimated for each layer and then incorporated in the calibration curve to improve the coefficient of determination. A field calibration showed an improvement of the determination coefficient by introducing bulk density and gypsum fraction, in addition to count ratio using moisture content on a volume basis as a dependent variable in multi linear regression analysis. The same procedure was successful with variable gravel fractions. (author). 18 refs., 3 figs., 2 tabs

  7. Calibration and use of filter test facility orifice plates

    Science.gov (United States)

    Fain, D. E.; Selby, T. W.

    1984-07-01

    There are three official DOE filter test facilities. These test facilities are used by the DOE, and others, to test nuclear grade HEPA filters to provide Quality Assurance that the filters meet the required specifications. The filters are tested for both filter efficiency and pressure drop. In the test equipment, standard orifice plates are used to set the specified flow rates for the tests. There has existed a need to calibrate the orifice plates from the three facilities with a common calibration source to assure that the facilities have comparable tests. A project has been undertaken to calibrate these orifice plates. In addition to reporting the results of the calibrations of the orifice plates, the means for using the calibration results will be discussed. A comparison of the orifice discharge coefficients for the orifice plates used at the seven facilities will be given. The pros and cons for the use of mass flow or volume flow rates for testing will be discussed. It is recommended that volume flow rates be used as a more practical and comparable means of testing filters. The rationale for this recommendation will be discussed.

  8. Calibration of sources for alpha spectroscopy systems

    International Nuclear Information System (INIS)

    Freitas, I.S.M.; Goncalez, O.L.

    1992-01-01

    This paper describes the calibration methodology for measuring the total alpha activity of plane and thin sources with the Alpha Spectrometer for Silicon Detector in the Nuclear Measures and Dosimetry laboratory at IEAv/CTA. (author)

  9. Control volume based hydrocephalus research; a phantom study

    Science.gov (United States)

    Cohen, Benjamin; Voorhees, Abram; Madsen, Joseph; Wei, Timothy

    2009-11-01

    Hydrocephalus is a complex spectrum of neurophysiological disorders involving perturbation of the intracranial contents; primarily increased intraventricular cerebrospinal fluid (CSF) volume and intracranial pressure are observed. CSF dynamics are highly coupled to the cerebral blood flows and pressures as well as the mechanical properties of the brain. Hydrocephalus, as such, is a very complex biological problem. We propose integral control volume analysis as a method of tracking these important interactions using mass and momentum conservation principles. As a first step in applying this methodology in humans, an in vitro phantom is used as a simplified model of the intracranial space. The phantom's design consists of a rigid container filled with a compressible gel. Within the gel a hollow spherical cavity represents the ventricular system and a cylindrical passage represents the spinal canal. A computer controlled piston pump supplies sinusoidal volume fluctuations into and out of the flow phantom. MRI is used to measure fluid velocity and volume change as functions of time. Independent pressure measurements and momentum flow rate measurements are used to calibrate the MRI data. These data are used as a framework for future work with live patients and normal individuals. Flow and pressure measurements on the flow phantom will be presented through the control volume framework.

  10. Variance of measurements from a calibration function derived from data which exhibit run-to-run differences

    International Nuclear Information System (INIS)

    Liebetrau, A.M.

    1985-01-01

    The volume of liquid in a nuclear process tank is determined from a calibration equation which expresses volume as a function of liquid level. Successive calibration runs are made to obtain data from which to estimate either the calibration function or its inverse. For tanks equipped with high-precision measurement systems to determine liquid level, it frequently happens that run-to-run differences due to uncontrolled or uncontrollable ambient conditions are large relative to within-run measurement errors. In the strict sense, a calibration function cannot be developed from data which exhibit significant run-to-run differences. In practice, run-to-run differences are ignored when they are small relative to the accuracy required for measurements of the tank's contents. The use of standard statistical techniques in this situation can result in variance estimates which severely underestimate the actual uncertainty in volume measurements. This paper gives a method whereby reasonable estimates of the calibration uncertainty in volume determinations can be obtained in the presence of statistically significant run-to-run variability. 4 references, 3 figures, 1 table

  11. Initial tank calibration at NUCEF critical facility. 2

    International Nuclear Information System (INIS)

    Yanagisawa, Hiroshi

    1994-07-01

    Analyses on initial tank calibration data were carried out for the purpose of the nuclear material accountancy and control for critical facilities in NUCEF: Nuclear Fuel Cycle Safety Engineering Research Facility. Calibration functions to evaluate volume of nuclear material solution in accountancy tanks were determined by regression analysis on the data considering dimension and shape of the tank. The analyses on dip-tube separation (probe separation), which are necessary to evaluate solution density in the tanks, were also carried out. As a result, regression errors of volume calculated with the calibration functions were within 0.05 lit. (0.01%) at a nominal level of Pu accountancy tanks. Errors of the evaluated dip-tube separations were also small, e.g. within 0.2mm (0.11%). Therefore, it was estimated that systematic errors of bulk measurements would satisfy the target value of NUCEF critical facilities (0.3% for Pu accountancy tanks). This paper summarizes the data analysis methods, results of analysis and evaluated errors. (author)

  12. Methodological issues in radiation dose-volume outcome analyses: Summary of a joint AAPM/NIH workshop

    International Nuclear Information System (INIS)

    Deasy, Joseph O.; Niemierko, Andrzej; Herbert, Donald; Yan, Di; Jackson, Andrew; Ten Haken, Randall K.; Langer, Mark; Sapareto, Steve

    2002-01-01

    This report represents a summary of presentations at a joint workshop of the National Institutes of Health and the American Association of Physicists in Medicine (AAPM). Current methodological issues in dose-volume modeling are addressed here from several different perspectives. Areas of emphasis include (a) basic modeling issues including the equivalent uniform dose framework and the bootstrap method, (b) issues in the valid use of statistics, including the need for meta-analysis, (c) issues in dealing with organ deformation and its effects on treatment response, (d) evidence for volume effects for rectal complications, (e) the use of volume effect data in liver and lung as a basis for dose escalation studies, and (f) implications of uncertainties in volume effect knowledge on optimized treatment planning. Taken together, these approaches to studying volume effects describe many implications for the development and use of this information in radiation oncology practice. Areas of significant interest for further research include the meta-analysis of clinical data; interinstitutional pooled data analyses of volume effects; analyses of the uncertainties in outcome prediction models, minimal parameter number outcome models for ranking treatment plans (e.g., equivalent uniform dose); incorporation of the effect of motion in the outcome prediction; dose-escalation/isorisk protocols based on outcome models; the use of functional imaging to study radio-response; and the need for further small animal tumor control probability/normal tissue complication probability studies

  13. SU-F-T-492: The Impact of Water Temperature On Absolute Dose Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Islam, N [State University of New York at Buffalo, Buffalo, NY (United States); Podgorsak, M [State University of New York at Buffalo, Buffalo, NY (United States); Roswell Park Cancer Institute, Buffalo, NY (United States)

    2016-06-15

    Purpose: The Task Group 51 (TG 51) protocol prescribes that dose calibration of photon beams be done by irradiating an ionization chamber in a water tank at pre-defined depths. Methodologies are provided to account for variations in measurement conditions by applying correction factors. However, the protocol does not completely account for the impact of water temperature. It is well established that water temperature will influence the density of air in the ion chamber collecting volume. Water temperature, however, will also influence the size of the collecting volume via thermal expansion of the cavity wall and the density of the water in the tank. In this work the overall effect of water temperature on absolute dosimetry has been investigated. Methods: Dose measurements were made using a Farmer-type ion chamber for 6 and 23 MV photon beams with water temperatures ranging from 10 to 40°C. A reference ion chamber was used to account for fluctuations in beam output between successive measurements. Results: For the same beam output, the dose determined using TG 51 was dependent on the temperature of the water in the tank. A linear regression of the data suggests that the dependence is statistically significant with p-values of the slope equal to 0.003 and 0.01 for 6 and 23 MV beams, respectively. For a 10 degree increase in water phantom temperature, the absolute dose determined with TG 51 increased by 0.27% and 0.31% for 6 and 23 MV beams, respectively. Conclusion: There is a measurable effect of water temperature on absolute dose calibration. To account for this effect, a reference temperature can be defined and a correction factor applied to account for deviations from this reference temperature during beam calibration. Such a factor is expected to be of similar magnitude to most of the existing TG 51 correction factors.

  14. SU-F-T-492: The Impact of Water Temperature On Absolute Dose Calibration

    International Nuclear Information System (INIS)

    Islam, N; Podgorsak, M

    2016-01-01

    Purpose: The Task Group 51 (TG 51) protocol prescribes that dose calibration of photon beams be done by irradiating an ionization chamber in a water tank at pre-defined depths. Methodologies are provided to account for variations in measurement conditions by applying correction factors. However, the protocol does not completely account for the impact of water temperature. It is well established that water temperature will influence the density of air in the ion chamber collecting volume. Water temperature, however, will also influence the size of the collecting volume via thermal expansion of the cavity wall and the density of the water in the tank. In this work the overall effect of water temperature on absolute dosimetry has been investigated. Methods: Dose measurements were made using a Farmer-type ion chamber for 6 and 23 MV photon beams with water temperatures ranging from 10 to 40°C. A reference ion chamber was used to account for fluctuations in beam output between successive measurements. Results: For the same beam output, the dose determined using TG 51 was dependent on the temperature of the water in the tank. A linear regression of the data suggests that the dependence is statistically significant with p-values of the slope equal to 0.003 and 0.01 for 6 and 23 MV beams, respectively. For a 10 degree increase in water phantom temperature, the absolute dose determined with TG 51 increased by 0.27% and 0.31% for 6 and 23 MV beams, respectively. Conclusion: There is a measurable effect of water temperature on absolute dose calibration. To account for this effect, a reference temperature can be defined and a correction factor applied to account for deviations from this reference temperature during beam calibration. Such a factor is expected to be of similar magnitude to most of the existing TG 51 correction factors.

  15. Safety assessment methodologies for near surface disposal facilities. Results of a co-ordinated research project (ISAM). Volume 1: Review and enhancement of safety assessment approaches and tools. Volume 2: Test cases

    International Nuclear Information System (INIS)

    2004-07-01

    the Safety Guide on 'Safety Assessment for Near Surface Disposal of Radioactive Waste' (Safety Standards Series No. WS-G- 1.1). The report of this CRP is presented in two volumes; Volume 1 contains a summary and a complete description of the ISAM project methodology and Volume 2 presents the application of the methodology to three hypothetical test cases

  16. One step geometrical calibration method for optical coherence tomography

    International Nuclear Information System (INIS)

    Díaz, Jesús Díaz; Ortmaier, Tobias; Stritzel, Jenny; Rahlves, Maik; Reithmeier, Eduard; Roth, Bernhard; Majdani, Omid

    2016-01-01

    We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery. (paper)

  17. Calibration of a Modified Californium Shuffler

    International Nuclear Information System (INIS)

    Sadowski, E.T.; Armstrong, F.; Oldham, R.; Ceo, R.; Williams, N.

    1995-01-01

    A californium shuffler originally designed to assay hollow cylindrical pieces of UA1 has been modified to assay solid cylinders. Calibration standards were characterized via chemical analysis of the molten UA1 taken during casting of the standards. The melt samples yielded much more reliable characterization data than drill samples taken from standards after the standards had solidified. By normalizing one well-characterized calibration curve to several standards at different enrichments, a relatively small number of standards was required to develop an enrichment-dependent calibration. The precision of this shuffler is 0.65%, and the typical random and systematic uncertainties are 0.53% and 0.73%, respectively, for a six minute assay of an ingot containing approximately 700 grams of 235 U. This paper will discuss (1) the discrepancies encountered when UA1 standards were characterized via melt samples versus drill samples, (2) a calibration methodology employing a small number of standards, and (3) a comparison of results from a previously unused shuffler with an existing shuffler. A small number of UA1 standards have been characterized using samples from the homogeneous molten state and have yielded enrichment-dependent and enrichment-independent calibration curves on two different shufflers

  18. Assessment methodology for new cooling lakes. Volume 3. Limnological and fisheries data and bibliography. Final report

    International Nuclear Information System (INIS)

    1981-10-01

    This is the data volume of the report entitled Assessment Methodology for New Cooling Lakes. Limnological and fisheries data were compiled in this volume for potential users in the utility industry. Published papers, reports, other written information, computer files, and direct contracts were used to compile a matrix of information. This volume presents data and the bibliographic sources of the power plant and geographical, limnological, and fisheries information for 181 lakes and reservoirs, of which 134 were used for cooling purposes. Data for 65 lakes were completed with respect to the limnology and fisheries parameters so that complete statistical analysis could be performed. Of these 65 lakes, 42 are used for cooling. Tables in this report contain data arranged by utility, power plant, limnology, water quality, morphoedaphic, and fishery categories. The data in the tables are keyed to a lake code. The references for the data shown are keyed to a numerical listing of the bibliography. Author, state, lake, and subject indexes facilitate searching for bibliographic information

  19. SCALA: In situ calibration for integral field spectrographs

    Science.gov (United States)

    Lombardo, S.; Küsters, D.; Kowalski, M.; Aldering, G.; Antilogus, P.; Bailey, S.; Baltay, C.; Barbary, K.; Baugh, D.; Bongard, S.; Boone, K.; Buton, C.; Chen, J.; Chotard, N.; Copin, Y.; Dixon, S.; Fagrelius, P.; Feindt, U.; Fouchez, D.; Gangler, E.; Hayden, B.; Hillebrandt, W.; Hoffmann, A.; Kim, A. G.; Leget, P.-F.; McKay, L.; Nordin, J.; Pain, R.; Pécontal, E.; Pereira, R.; Perlmutter, S.; Rabinowitz, D.; Reif, K.; Rigault, M.; Rubin, D.; Runge, K.; Saunders, C.; Smadja, G.; Suzuki, N.; Taubenberger, S.; Tao, C.; Thomas, R. C.; Nearby Supernova Factory

    2017-11-01

    Aims: The scientific yield of current and future optical surveys is increasingly limited by systematic uncertainties in the flux calibration. This is the case for type Ia supernova (SN Ia) cosmology programs, where an improved calibration directly translates into improved cosmological constraints. Current methodology rests on models of stars. Here we aim to obtain flux calibration that is traceable to state-of-the-art detector-based calibration. Methods: We present the SNIFS Calibration Apparatus (SCALA), a color (relative) flux calibration system developed for the SuperNova integral field spectrograph (SNIFS), operating at the University of Hawaii 2.2 m (UH 88) telescope. Results: By comparing the color trend of the illumination generated by SCALA during two commissioning runs, and to previous laboratory measurements, we show that we can determine the light emitted by SCALA with a long-term repeatability better than 1%. We describe the calibration procedure necessary to control for system aging. We present measurements of the SNIFS throughput as estimated by SCALA observations. Conclusions: The SCALA calibration unit is now fully deployed at the UH 88 telescope, and with it color-calibration between 4000 Å and 9000 Å is stable at the percent level over a one-year baseline.

  20. A New Calibration Methodology for Thorax and Upper Limbs Motion Capture in Children Using Magneto and Inertial Sensors

    Directory of Open Access Journals (Sweden)

    Luca Ricci

    2014-01-01

    Full Text Available Recent advances in wearable sensor technologies for motion capture have produced devices, mainly based on magneto and inertial measurement units (M-IMU, that are now suitable for out-of-the-lab use with children. In fact, the reduced size, weight and the wireless connectivity meet the requirement of minimum obtrusivity and give scientists the possibility to analyze children’s motion in daily life contexts. Typical use of magneto and inertial measurement units (M-IMU motion capture systems is based on attaching a sensing unit to each body segment of interest. The correct use of this setup requires a specific calibration methodology that allows mapping measurements from the sensors’ frames of reference into useful kinematic information in the human limbs’ frames of reference. The present work addresses this specific issue, presenting a calibration protocol to capture the kinematics of the upper limbs and thorax in typically developing (TD children. The proposed method allows the construction, on each body segment, of a meaningful system of coordinates that are representative of real physiological motions and that are referred to as functional frames (FFs. We will also present a novel cost function for the Levenberg–Marquardt algorithm, to retrieve the rotation matrices between each sensor frame (SF and the corresponding FF. Reported results on a group of 40 children suggest that the method is repeatable and reliable, opening the way to the extensive use of this technology for out-of-the-lab motion capture in children.

  1. The S-NPP VIIRS Day-Night Band On-Orbit Calibration/Characterization and Current State of SDR Products

    Directory of Open Access Journals (Sweden)

    Shihyan Lee

    2014-12-01

    Full Text Available The launch of VIIRS on-board the Suomi-National Polar-orbiting Partnership (S-NPP on 28 October 2011, marked the beginning of the next chapter on nighttime lights observation started by the Defense Meteorological Satellite Program’s (DMSP OLS sensor more than two decades ago. The VIIRS observes the nighttime lights on Earth through its day-night band (DNB, a panchromatic channel covering the wavelengths from 500 nm to 900 nm. Compared to its predecessors, the VIIRS DNB has a much improved spatial/temporal resolution, radiometric sensitivity and, more importantly, continuous calibration using on-board calibrators (OBCs. In this paper, we describe the current state of the NASA calibration and characterization methodology used in supporting mission data quality assurance and producing consistent mission-wide sensor data records (SDRs through NASA’s Land Product Evaluation and Analysis Tool Element (Land PEATE. The NASA calibration method utilizes the OBCs to determine gains, offset drift and sign-to-noise ratio (SNR over the entire mission. In gain determination, the time-dependent relative spectral response (RSR is used to correct the optical throughput change over time. A deep space view acquired during an S-NPP pitch maneuver is used to compute the airglow free dark offset for DNB’s high gain stage. The DNB stray light is estimated each month from new-moon dark Earth surface observations to remove the excessive stray light over the day-night terminators. As the VIIRS DNB on-orbit calibration is the first of its kind, the evolution of the calibration methodology is evident when the S-NPP VIIRS’s official calibrations are compared with our latest mission-wide reprocessing. In the future, the DNB calibration methodology is likely to continue evolving, and the mission-wide reprocessing is a key to providing consistently calibrated DNB SDRs for the user community. In the meantime, the NASA Land PEATE provides an alternative source to obtain

  2. Stochastic calibration and learning in nonstationary hydroeconomic models

    Science.gov (United States)

    Maneta, M. P.; Howitt, R.

    2014-05-01

    Concern about water scarcity and adverse climate events over agricultural regions has motivated a number of efforts to develop operational integrated hydroeconomic models to guide adaptation and optimal use of water. Once calibrated, these models are used for water management and analysis assuming they remain valid under future conditions. In this paper, we present and demonstrate a methodology that permits the recursive calibration of economic models of agricultural production from noisy but frequently available data. We use a standard economic calibration approach, namely positive mathematical programming, integrated in a data assimilation algorithm based on the ensemble Kalman filter equations to identify the economic model parameters. A moving average kernel ensures that new and past information on agricultural activity are blended during the calibration process, avoiding loss of information and overcalibration for the conditions of a single year. A regularization constraint akin to the standard Tikhonov regularization is included in the filter to ensure its stability even in the presence of parameters with low sensitivity to observations. The results show that the implementation of the PMP methodology within a data assimilation framework based on the enKF equations is an effective method to calibrate models of agricultural production even with noisy information. The recursive nature of the method incorporates new information as an added value to the known previous observations of agricultural activity without the need to store historical information. The robustness of the method opens the door to the use of new remote sensing algorithms for operational water management.

  3. Establishment of a new calibration method of pencil ionization chamber for dosimetry in computed tomography

    International Nuclear Information System (INIS)

    Dias, Daniel Menezes

    2010-01-01

    Pencil ionization chambers are used for beam dosimetry in computed tomography equipment (CT). In this study, a new calibration methodology was established, in order to make the Calibration Laboratory of Instituto de Pesquisas Energeticas e Nucleares (LCI) suitable to international metrological standards, dealing with specific procedures for calibration of these chambers used in CT. Firstly, the setup for the new RQT radiation qualities was mounted, in agreement with IEC61267 from the International Electrotechnical Commission (IEC). After the establishment of these radiation qualities, a specific calibration methodology for pencil ionization chambers was set, according to Technical Report Series No. 457, from the International Atomic Energy Agency (IAEA), which describes particularities of the procedure to be followed by the Secondary Standard Dosimetry Laboratories (SSDL's), concerning to collimation and positioning related to the radiation beam. Initially, PPV (kV) measurements and the determination of copper additional filtrations were carried out, measuring the half value layers (HVL) recommended by the IEC 61267 standard, after that the RQT 8, RQT 9 and RQT 10 radiation quality references were established. For additional filters, aluminum and copper of high purity (around 99.9%) were used. RQT's in thickness of copper filters equivalent to the set 'RQR (Al) + Additional Filtration (Cu)' was directly found by an alternative methodology used to determine additional filtrations, which is a good option when RQR's have not the possibility of be setting up. With the establishment of this new methodology for the ionization pencil chambers calibration, the LCI is ready to calibrate these instruments according to the most recent international standards. Therefore, an improvement in calibration traceability, as well as in metrological services offered by IPEN to all Brazil is achieved. (author)

  4. VIIRS reflective solar bands on-orbit calibration and performance: a three-year update

    Science.gov (United States)

    Sun, Junqiang; Wang, Menghua

    2014-11-01

    The on-orbit calibration of the reflective solar bands (RSBs) of VIIRS and the result from the analysis of the up-to-date 3 years of mission data are presented. The VIIRS solar diffuser (SD) and lunar calibration methodology are discussed, and the calibration coefficients, called F-factors, for the RSBs are given for the latest reincarnation. The coefficients derived from the two calibrations are compared and the uncertainties of the calibrations are discussed. Numerous improvements are made, with the major improvement to the calibration result come mainly from the improved bidirectional reflectance factor (BRF) of the SD and the vignetting functions of both the SD screen and the sun-view screen. The very clean results, devoid of many previously known noises and artifacts, assures that VIIRS has performed well for the three years on orbit since launch, and in particular that the solar diffuser stability monitor (SDSM) is functioning essentially without flaws. The SD degradation, or H-factors, for most part shows the expected decline except for the surprising rise on day 830 lasting for 75 days signaling a new degradation phenomenon. Nevertheless the SDSM and the calibration methodology have successfully captured the SD degradation for RSB calibration. The overall improvement has the most significant and direct impact on the ocean color products which demands high accuracy from RSB observations.

  5. Calibration of CR-39 for radon-related parameters using sealed cup technique

    International Nuclear Information System (INIS)

    Abo-Elmagd, M.; Daif, M. M.

    2010-01-01

    Effective radium content, mass and areal radon exhalation rates of soil and rock samples are important radon-related parameters and can be used as a better indicator of radon risk. A sealed cup fitted to a CR-39 detector and to the sample under measurement is an advantageous passive device for the measurement of these parameters. The main factors affecting the results are the detector calibration factor and the sample weight. The results of an active technique (Lucas cell) and the CR-39 detector have been found to be correlated resulting in a reliable detector calibration factor. The result illustrates the dependence of the CR-39 calibration factor with the sample weight which is difficult to use in practice, because each sample weight has its own calibration factor of CR-39. It is reported to demonstrate the advantage of a back diffusion correction. After correcting the results for back diffusion effects, one obtains an approximately constant calibration factor for the sample volumes up to one-third the total sealed cup volume. For this condition the calibration factor is equal to 0.237 track cm -2 per Bq m -3 d with about 1% uncertainty. (authors)

  6. Thermal power calibrations of the IPR-R1 TRIGA reactor by the calorimetric and the heat balance methods

    International Nuclear Information System (INIS)

    Mesquita, Amir Zacarias; Rezende, Hugo Cesar; Souza, Rose Mary Gomes do Prado

    2009-01-01

    Since the first nuclear reactor was built, a number of methodological variations have been evolved for the calibration of the reactor thermal power. Power monitoring of reactors is done by means of neutronic instruments, but its calibration is always done by thermal procedures. The purpose of this paper is to present the results of the thermal power calibration carried out on March 5th, 2009 in the IPR-R1 TRIGA reactor. It was used two procedures: the calorimetric and heat balance methods. The calorimetric procedure was done with the reactor operating at a constant power, with primary cooling system switched off. The rate of temperature rise of the water was recorded. The reactor power is calculate as a function of the temperature-rise rate and the system heat capacity constant. The heat balance procedure consists in the steady-state energy balance of the primary cooling loop of the reactor. For this balance, the inlet and outlet temperatures and the water flow in the primary cooling loop were measured. The heat transferred through the primary loop was added to the heat leakage from the reactor pool. The calorimetric method calibration presented a large uncertainty. The main source of error was the determination of the heat content of the system, due to a large uncertainty in the volume of the water in the system and a lack of homogenization of the water temperature. The heat balance calibration in the primary loop is the standard procedure for calibrating the power of the IPR-R1 TRIGA nuclear reactor. (author))

  7. Calibration of the NPL secondary standard radionuclide calibrator for the new 10R Schott, Type 1+ vials

    International Nuclear Information System (INIS)

    Baker, M.

    2005-01-01

    For many years, P6 vials have been used for the distribution of a wide range of diagnostic and therapeutic radioactive solutions. The activity measurements were performed in this geometry and, in time, the UK calibration system for nuclear medicine was based on this container as a standard. However, one major supplier of radiopharmaceuticals has replaced the P6 vial with the 10R Type 1+ Schott vial. As the dimensions of the new vial are different from those of the P6 vial and the responses of radionuclide calibrators are known to be container dependent, the need for re-calibration became apparent. Preliminary measurements made on some typical radionuclide calibrators for 125 I solution indicated a difference in response of about 10% between the two vials. The master ionisation chamber of the NPL secondary standard radionuclide calibrator has been re-calibrated and new calibration factors and volume correction factors for 10R Schott vials have been derived for the relevant medical radionuclides. The standard holder was also modified to accommodate the new larger vial. The complete list of factors and the method used to determine them is presented in this paper. The availability of these new factors will improve the quality of activity measurements in nuclear medicine, as calibration services can now be provided by NPL for the new container. These factors can also be employed for all commercial NPL secondary standard radionuclide calibrators (now known as the NPL-CRC and previously as the 671 or ISOCAL IV)

  8. Intravascular volume in cirrhosis. Reassessment using improved methodology

    International Nuclear Information System (INIS)

    Rector, W.G. Jr.; Ibarra, F.

    1988-01-01

    Previous studies of blood volume (BV) in cirrhosis have either not adjusted BV properly for body size; determined plasma volume from the dilution of labeled albumin 10-20 min postinjection, when some extravascular redistribution has already occurred; and/or not used the correct whole body-peripheral hematocrit ratio (0.82) in calculating whole BV from plasma volume and the peripheral hematocrit. We measured BV with attention to these considerations in 19 patients with cirrhosis and reexamined the determinants of vascular volume and the relationship between vascular volume and sodium retention. BV was calculated as plasma volume (determined from extrapolated plasma activity of intravenously injected [ 131 I]+albumin at time 0) divided by (peripheral hematocrit X 0.82). The result was expressed per kilogram dry body weight, determined by subtracting the mass of ascites (measured by isotope dilution; 1 liter = 1 kg) from the actual body weight of nonedematous patients. Measured and expressed in this way, BV correlated strongly with esophageal variceal size (r = 0.87, P less than 0.05), although not with net portal, right atrial, inferior vena caval, or arterial pressure, and was significantly greater in patients with sodium retention as compared to patients without sodium retention. The principal modifier of vascular volume in cirrhosis is vascular capacity, which is probably mainly determined by the extent of the portasystemic collateral circulation. Increased vascular volume in patients with sodium retention as compared to patients without sodium retention supports the overflow theory of ascites formation

  9. Comparison of infusion pumps calibration methods

    Science.gov (United States)

    Batista, Elsa; Godinho, Isabel; do Céu Ferreira, Maria; Furtado, Andreia; Lucas, Peter; Silva, Claudia

    2017-12-01

    Nowadays, several types of infusion pump are commonly used for drug delivery, such as syringe pumps and peristaltic pumps. These instruments present different measuring features and capacities according to their use and therapeutic application. In order to ensure the metrological traceability of these flow and volume measuring equipment, it is necessary to use suitable calibration methods and standards. Two different calibration methods can be used to determine the flow error of infusion pumps. One is the gravimetric method, considered as a primary method, commonly used by National Metrology Institutes. The other calibration method, a secondary method, relies on an infusion device analyser (IDA) and is typically used by hospital maintenance offices. The suitability of the IDA calibration method was assessed by testing several infusion instruments at different flow rates using the gravimetric method. In addition, a measurement comparison between Portuguese Accredited Laboratories and hospital maintenance offices was performed under the coordination of the Portuguese Institute for Quality, the National Metrology Institute. The obtained results were directly related to the used calibration method and are presented in this paper. This work has been developed in the framework of the EURAMET projects EMRP MeDD and EMPIR 15SIP03.

  10. Calibration procedure for Slocum glider deployed optical instruments.

    Science.gov (United States)

    Cetinić, Ivona; Toro-Farmer, Gerardo; Ragan, Matthew; Oberg, Carl; Jones, Burton H

    2009-08-31

    Recent developments in the field of the autonomous underwater vehicles allow the wide usage of these platforms as part of scientific experiments, monitoring campaigns and more. The vehicles are often equipped with sensors measuring temperature, conductivity, chlorophyll a fluorescence (Chl a), colored dissolved organic matter (CDOM) fluorescence, phycoerithrin (PE) fluorescence and spectral volume scattering function at 117 degrees, providing users with high resolution, real time data. However, calibration of these instruments can be problematic. Most in situ calibrations are performed by deploying complementary instrument packages or water samplers in the proximity of the glider. Laboratory calibrations of the mounted sensors are difficult due to the placement of the instruments within the body of the vehicle. For the laboratory calibrations of the Slocum glider instruments we developed a small calibration chamber where we can perform precise calibrations of the optical instruments aboard our glider, as well as sensors from other deployment platforms. These procedures enable us to obtain pre- and post-deployment calibrations for optical fluorescence instruments, which may differ due to the biofouling and other physical damage that can occur during long-term glider deployments. We found that biofouling caused significant changes in the calibration scaling factors of fluorescent sensors, suggesting the need for consistent and repetitive calibrations for gliders as proposed in this paper.

  11. A laser calibration system for the STAR TPC

    CERN Document Server

    Lebedev, A

    2002-01-01

    A Time Projection Chamber (TPC) is the primary tracking detector for the STAR experiment at RHIC. A laser calibration system was built to calibrate and monitor the TPC tracking performance. The laser system uses a novel design which produces approx 500 thin, ionizing beams distributed throughout the tracking volume. This new approach is significantly simpler than the traditional ones, and provides complete TPC coverage at a reduced cost. The laser system was used during the RHIC 2000 summer run to measure drift velocities with about 0.02% accuracy and to monitor the TPC performance. Calibration runs were made with and without a magnetic field to check B field map corrections.

  12. Methodology to include a correction for offset in the calibration of a Diode-based 2D verification device; Metodologia para incluir una correccion por offset en la calibracion de un dispositivo de verificacion 2D basado en diodos

    Energy Technology Data Exchange (ETDEWEB)

    Ramirez Ros, J. C.; Pamos Urena, M.; Jerez Sainz, M.; Lobato Munoz, M.; Jodar Lopez, C. A.; Ruiz Lopez, M. a.; Carrasco Rodriguez, J. L.

    2013-07-01

    We propose a methodology to correct doses of device verification 2D MapChek2 planes by offset. This methodology provides an array of correction by Offset applied to the calibration per dose due to the Offset of the diode Central as well as the correction of the Offset of each diode on each acquisition. (Author)

  13. The Utilization of Background Spectrum to Calibrate Gamma Spectrometry

    International Nuclear Information System (INIS)

    Mahrouka, M. M.; Mutawa, A. M.

    2004-01-01

    Many developed countries have very poor reference standards to calibrate their nuclear instrumentations or may find some difficulties to obtain a reference standard. In this work a simple way for Gamma spectrometry calibration was developed. The method depends on one reference point and additional points from the background. The two derived equations were applied to the analyses of radioactive nuclides in soil and liquid samples prepared by IAEA laboratories through AL MERA Project. The results showed the precision of the methodology used, as well as, the possibility of using some points in the background spectrum as a replacement for reference standards of Gamma spectrometry calibration. (authors)

  14. Horizontal scale calibration of theodolites and total station using a gauge index table

    International Nuclear Information System (INIS)

    Vieira, L H B; Filho, W L O; Barros, W S

    2015-01-01

    This paper shows a methodology to calibrate the horizontal scale of theodolites and total station using a high accuracy index table. The calibration pursued the method of circular scales and precision polygons (also called Rosette Method [1] or multistep). This method consists in the angle comparison of two circular divisions in all relative positions possibilities. Index table errors and theodolite horizontal scale errors were obtained using the method of least squares which is used to process the data from Rosette Method. An experimental setup was used to evaluate this methodology and the details of the mechanical assembly are also described in this paper. Several theodolites and total stations were calibrated using the proposed system and the results infer that the method is suitable to calibrate the different models available in the market. The system showed good stability over time with measurements uncertainties around 1' (one second) depending on instrument features. (paper)

  15. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-06-02

    This study addresses the effect of calibration methodologies on calibration responsivities and the resulting impact on radiometric measurements. The calibration responsivities used in this study are provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The BORCAL method provides outdoor calibration responsivity of pyranometers and pyrheliometers at a 45 degree solar zenith angle and responsivity as a function of solar zenith angle determined by clear-sky comparisons to reference irradiance. The BORCAL method also employs a thermal offset correction to the calibration responsivity of single-black thermopile detectors used in pyranometers. Indoor calibrations of radiometers by their manufacturers are performed using a stable artificial light source in a side-by-side comparison of the test radiometer under calibration to a reference radiometer of the same type. These different methods of calibration demonstrated 1percent to 2 percent differences in solar irradiance measurement. Analyzing these values will ultimately enable a reduction in radiometric measurement uncertainties and assist in developing consensus on a standard for calibration.

  16. The Efficiency of OLS Estimators of Structural Parameters in a Simple Linear Regression Model in the Calibration of the Averages Scheme

    Directory of Open Access Journals (Sweden)

    Kowal Robert

    2016-12-01

    Full Text Available A simple linear regression model is one of the pillars of classic econometrics. Multiple areas of research function within its scope. One of the many fundamental questions in the model concerns proving the efficiency of the most commonly used OLS estimators and examining their properties. In the literature of the subject one can find taking back to this scope and certain solutions in that regard. Methodically, they are borrowed from the multiple regression model or also from a boundary partial model. Not everything, however, is here complete and consistent. In the paper a completely new scheme is proposed, based on the implementation of the Cauchy-Schwarz inequality in the arrangement of the constraint aggregated from calibrated appropriately secondary constraints of unbiasedness which in a result of choice the appropriate calibrator for each variable directly leads to showing this property. A separate range-is a matter of choice of such a calibrator. These deliberations, on account of the volume and kinds of the calibration, were divided into a few parts. In the one the efficiency of OLS estimators is proven in a mixed scheme of the calibration by averages, that is preliminary, and in the most basic frames of the proposed methodology. In these frames the future outlines and general premises constituting the base of more distant generalizations are being created.

  17. Evaluation of the energy dependence of ionization chambers pencil type calibrated beam tomography standards

    International Nuclear Information System (INIS)

    Fontes, Ladyjane Pereira; Potiens, Maria da Penha A.

    2015-01-01

    The Instrument Calibration Laboratory of IPEN (LCI - IPEN) performs calibrations of pencil-type ionization chambers (IC) used in measures of dosimetric survey on clinical systems of Computed Tomography (CT). Many users make mistakes when using a calibrated ionization chamber in their CT dosimetry systems. In this work a methodology for determination of factors of correction for quality (Kq) through the calibration curve that is specific for each ionization chamber was established. Furthermore, it was possible to demonstrate the energy dependence on an pencil-type Ionization Chamber(IC) calibrated at the LCI - IPEN. (author)

  18. [Fundamental aspects for accrediting medical equipment calibration laboratories in Colombia].

    Science.gov (United States)

    Llamosa-Rincón, Luis E; López-Isaza, Giovanni A; Villarreal-Castro, Milton F

    2010-02-01

    Analysing the fundamental methodological aspects which should be considered when drawing up calibration procedure for electro-medical equipment, thereby permitting international standard-based accreditation of electro-medical metrology laboratories in Colombia. NTC-ISO-IEC 17025:2005 and GTC-51-based procedures for calibrating electro-medical equipment were implemented and then used as patterns. The mathematical model for determining the estimated uncertainty value when calibrating electro-medical equipment for accreditation by the Electrical Variable Metrology Laboratory's Electro-medical Equipment Calibration Area accredited in compliance with Superintendence of Industry and Commerce Resolution 25771 May 26th 2009 consists of two equations depending on the case; they are: E = (Ai + sigmaAi) - (Ar + sigmaAr + deltaAr1) and E = (Ai + sigmaAi) - (Ar + sigmaA + deltaAr1). The mathematical modelling implemented for measuring uncertainty in the Universidad Tecnológica de Pereira's Electrical Variable Metrology Laboratory (Electro-medical Equipment Calibration Area) will become a good guide for calibration initiated in other laboratories in Colombia and Latin-America.

  19. Exercise of laboratory comparison for contamination monitor calibration between LNMRI/IRD and LCR/UERJ - 2016

    International Nuclear Information System (INIS)

    Cabral, T.S.; David, M.

    2016-01-01

    This work was motivated by the need to decide on the best methodology to be applied in the next contamination monitor calibration comparisons with the Brazilian network of calibration radiation monitors. The calibration factor was chosen as a response calibration performed in the four monitors used in this comparison because it does not require the detector area or probe thereby reducing an important variable. It was observed that the variation of the positioning system may have an influence up to 10% in calibration. The results obtained for the calibration factor showed a difference of up to 31.2%. (author)

  20. Calibration factor or calibration coefficient?

    International Nuclear Information System (INIS)

    Meghzifene, A.; Shortt, K.R.

    2002-01-01

    Full text: The IAEA/WHO network of SSDLs was set up in order to establish links between SSDL members and the international measurement system. At the end of 2001, there were 73 network members in 63 Member States. The SSDL network members provide calibration services to end-users at the national or regional level. The results of the calibrations are summarized in a document called calibration report or calibration certificate. The IAEA has been using the term calibration certificate and will continue using the same terminology. The most important information in a calibration certificate is a list of calibration factors and their related uncertainties that apply to the calibrated instrument for the well-defined irradiation and ambient conditions. The IAEA has recently decided to change the term calibration factor to calibration coefficient, to be fully in line with ISO [ISO 31-0], which recommends the use of the term coefficient when it links two quantities A and B (equation 1) that have different dimensions. The term factor should only be used for k when it is used to link the terms A and B that have the same dimensions A=k.B. However, in a typical calibration, an ion chamber is calibrated in terms of a physical quantity such as air kerma, dose to water, ambient dose equivalent, etc. If the chamber is calibrated together with its electrometer, then the calibration refers to the physical quantity to be measured per electrometer unit reading. In this case, the terms referred have different dimensions. The adoption by the Agency of the term coefficient to express the results of calibrations is consistent with the 'International vocabulary of basic and general terms in metrology' prepared jointly by the BIPM, IEC, ISO, OIML and other organizations. The BIPM has changed from factor to coefficient. The authors believe that this is more than just a matter of semantics and recommend that the SSDL network members adopt this change in terminology. (author)

  1. A single model procedure for tank calibration function estimation

    International Nuclear Information System (INIS)

    York, J.C.; Liebetrau, A.M.

    1995-01-01

    Reliable tank calibrations are a vital component of any measurement control and accountability program for bulk materials in a nuclear reprocessing facility. Tank volume calibration functions used in nuclear materials safeguards and accountability programs are typically constructed from several segments, each of which is estimated independently. Ideally, the segments correspond to structural features in the tank. In this paper the authors use an extension of the Thomas-Liebetrau model to estimate the entire calibration function in a single step. This procedure automatically takes significant run-to-run differences into account and yields an estimate of the entire calibration function in one operation. As with other procedures, the first step is to define suitable calibration segments. Next, a polynomial of low degree is specified for each segment. In contrast with the conventional practice of constructing a separate model for each segment, this information is used to set up the design matrix for a single model that encompasses all of the calibration data. Estimation of the model parameters is then done using conventional statistical methods. The method described here has several advantages over traditional methods. First, modeled run-to-run differences can be taken into account automatically at the estimation step. Second, no interpolation is required between successive segments. Third, variance estimates are based on all the data, rather than that from a single segment, with the result that discontinuities in confidence intervals at segment boundaries are eliminated. Fourth, the restrictive assumption of the Thomas-Liebetrau method, that the measured volumes be the same for all runs, is not required. Finally, the proposed methods are readily implemented using standard statistical procedures and widely-used software packages

  2. Review of Calibration Methods for Scheimpflug Camera

    Directory of Open Access Journals (Sweden)

    Cong Sun

    2018-01-01

    Full Text Available The Scheimpflug camera offers a wide range of applications in the field of typical close-range photogrammetry, particle image velocity, and digital image correlation due to the fact that the depth-of-view of Scheimpflug camera can be greatly extended according to the Scheimpflug condition. Yet, the conventional calibration methods are not applicable in this case because the assumptions used by classical calibration methodologies are not valid anymore for cameras undergoing Scheimpflug condition. Therefore, various methods have been investigated to solve the problem over the last few years. However, no comprehensive review exists that provides an insight into recent calibration methods of Scheimpflug cameras. This paper presents a survey of recent calibration methods of Scheimpflug cameras with perspective lens, including the general nonparametric imaging model, and analyzes in detail the advantages and drawbacks of the mainstream calibration models with respect to each other. Real data experiments including calibrations, reconstructions, and measurements are performed to assess the performance of the models. The results reveal that the accuracies of the RMM, PLVM, PCIM, and GNIM are basically equal, while the accuracy of GNIM is slightly lower compared with the other three parametric models. Moreover, the experimental results reveal that the parameters of the tangential distortion are likely coupled with the tilt angle of the sensor in Scheimpflug calibration models. The work of this paper lays the foundation of further research of Scheimpflug cameras.

  3. Calibration, validation, and sensitivity analysis: What's what

    International Nuclear Information System (INIS)

    Trucano, T.G.; Swiler, L.P.; Igusa, T.; Oberkampf, W.L.; Pilch, M.

    2006-01-01

    One very simple interpretation of calibration is to adjust a set of parameters associated with a computational science and engineering code so that the model agreement is maximized with respect to a set of experimental data. One very simple interpretation of validation is to quantify our belief in the predictive capability of a computational code through comparison with a set of experimental data. Uncertainty in both the data and the code are important and must be mathematically understood to correctly perform both calibration and validation. Sensitivity analysis, being an important methodology in uncertainty analysis, is thus important to both calibration and validation. In this paper, we intend to clarify the language just used and express some opinions on the associated issues. We will endeavor to identify some technical challenges that must be resolved for successful validation of a predictive modeling capability. One of these challenges is a formal description of a 'model discrepancy' term. Another challenge revolves around the general adaptation of abstract learning theory as a formalism that potentially encompasses both calibration and validation in the face of model uncertainty

  4. A self-calibrating radon monitor with statistical discrimination

    International Nuclear Information System (INIS)

    Valcov, N.; Purghel, L.

    2002-01-01

    A radon monitor, able to perform the measurement of the radon and its progeny volumic activity, in a gamma-ray or natural radiation background field, was developed. The instrument consists of a 10 l ionization chamber, a high voltage source, an integrating preamplifier, a data acquisition system and a personal computer. A new method for self-calibration of Radon volumic activity measurements, based on the alpha counting with an ionization chamber is also presented

  5. Noninvasive method for the calibration of the peak voltage (kVp) meters

    International Nuclear Information System (INIS)

    Macedo, E.M.; Navarro, M.V.T.; Pereira, L.; Garcia, I.F.M.; Navarro, V.C.C.

    2015-01-01

    Quality control in diagnostic radiology is one of the mechanisms that minimize radiation exposure, and the measurement of tube voltage is one of the main test in these procedures. So, the calibration of non-invasive tube voltage meters is essential to maintain the metrological reliability of quality control tests. Thus, this work describes the implementation of the calibration methodology of the quantity tube peak voltage by the substitution method, using non-invasive standard meter, at LABPROSAUD-IFBA. The results showed great performance and when compared with calibrations by invasive methods, showed maximum difference of 4%, contemplated in the uncertainty ranges of the calibrations. (author)

  6. Calibration of in vitro bioassay methodology for determination “1”3”1I urine

    International Nuclear Information System (INIS)

    Carvalho, C.B.; Hazin, C.; Lima, A.R.

    2015-01-01

    The use of unsealed radioactive sources in institutions practicing nuclear medicine poses a significant risk of internal exposure of workers. In this context, handling of “1”3”1I plays an important role in relation to other radionuclides due to its wide application, particularly in medical diagnosis and therapy of diseases related to the thyroid gland. Given the increasing number of services using “1”3”1I in their examination protocols, the probability of accidental incorporation of this radionuclide has increased. The present study aimed to implement methodologies for in vitro bioassay at the Centro Regional de Ciências Nucleares do Nordeste (CRCN-NE/CNEN), Recife, Brazil, for internal monitoring of individuals occupationally exposed to “1”3”1I. For in vitro system calibration, a coaxial HPGe detector model GC1018 and a standard “1”3”3Ba source were used. Upon obtaining the calibration factor, it was possible to determine the minimum detectable activities (MDA) for the system by using direct measurements of distilled water simulating urine (in vitro). Then, by using the biokinetic models provided by the International Commission on Radiological Protection, edited with the AIDE software version 6.0, it was possible to estimate the Minimum Detectable Effective Dose (MDED). MDED values obtained were compared to the record level of 1 mSv recommended by the International Atomic Energy Agency in the urine compartment 24 h. The values found were lower than the record level of 1 mSv in all simulated incorporation scenarios: inhalation of vapor and particles with AMAD of 1 μ and 5 μ, type F compound, and ingestion. The results of this work show that the implemented technique is suitable for conducting internal monitoring of workers to “1”3”1I. It is intended to continue the work aiming the monitoring of occupationally exposed individuals from Nuclear Medicine Services in Recife, Brazil. (authors)

  7. Body composition in Nepalese children using isotope dilution: the production of ethnic-specific calibration equations and an exploration of methodological issues.

    Science.gov (United States)

    Devakumar, Delan; Grijalva-Eternod, Carlos S; Roberts, Sebastian; Chaube, Shiva Shankar; Saville, Naomi M; Manandhar, Dharma S; Costello, Anthony; Osrin, David; Wells, Jonathan C K

    2015-01-01

    Background. Body composition is important as a marker of both current and future health. Bioelectrical impedance (BIA) is a simple and accurate method for estimating body composition, but requires population-specific calibration equations. Objectives. (1) To generate population specific calibration equations to predict lean mass (LM) from BIA in Nepalese children aged 7-9 years. (2) To explore methodological changes that may extend the range and improve accuracy. Methods. BIA measurements were obtained from 102 Nepalese children (52 girls) using the Tanita BC-418. Isotope dilution with deuterium oxide was used to measure total body water and to estimate LM. Prediction equations for estimating LM from BIA data were developed using linear regression, and estimates were compared with those obtained from the Tanita system. We assessed the effects of flexing the arms of children to extend the range of coverage towards lower weights. We also estimated potential error if the number of children included in the study was reduced. Findings. Prediction equations were generated, incorporating height, impedance index, weight and sex as predictors (R (2) 93%). The Tanita system tended to under-estimate LM, with a mean error of 2.2%, but extending up to 25.8%. Flexing the arms to 90° increased the lower weight range, but produced a small error that was not significant when applied to children <16 kg (p 0.42). Reducing the number of children increased the error at the tails of the weight distribution. Conclusions. Population-specific isotope calibration of BIA for Nepalese children has high accuracy. Arm position is important and can be used to extend the range of low weight covered. Smaller samples reduce resource requirements, but leads to large errors at the tails of the weight distribution.

  8. Body composition in Nepalese children using isotope dilution: the production of ethnic-specific calibration equations and an exploration of methodological issues

    Directory of Open Access Journals (Sweden)

    Delan Devakumar

    2015-03-01

    Full Text Available Background. Body composition is important as a marker of both current and future health. Bioelectrical impedance (BIA is a simple and accurate method for estimating body composition, but requires population-specific calibration equations.Objectives. (1 To generate population specific calibration equations to predict lean mass (LM from BIA in Nepalese children aged 7–9 years. (2 To explore methodological changes that may extend the range and improve accuracy.Methods. BIA measurements were obtained from 102 Nepalese children (52 girls using the Tanita BC-418. Isotope dilution with deuterium oxide was used to measure total body water and to estimate LM. Prediction equations for estimating LM from BIA data were developed using linear regression, and estimates were compared with those obtained from the Tanita system. We assessed the effects of flexing the arms of children to extend the range of coverage towards lower weights. We also estimated potential error if the number of children included in the study was reduced.Findings. Prediction equations were generated, incorporating height, impedance index, weight and sex as predictors (R2 93%. The Tanita system tended to under-estimate LM, with a mean error of 2.2%, but extending up to 25.8%. Flexing the arms to 90° increased the lower weight range, but produced a small error that was not significant when applied to children <16 kg (p 0.42. Reducing the number of children increased the error at the tails of the weight distribution.Conclusions. Population-specific isotope calibration of BIA for Nepalese children has high accuracy. Arm position is important and can be used to extend the range of low weight covered. Smaller samples reduce resource requirements, but leads to large errors at the tails of the weight distribution.

  9. Gradient-based model calibration with proxy-model assistance

    Science.gov (United States)

    Burrows, Wesley; Doherty, John

    2016-02-01

    Use of a proxy model in gradient-based calibration and uncertainty analysis of a complex groundwater model with large run times and problematic numerical behaviour is described. The methodology is general, and can be used with models of all types. The proxy model is based on a series of analytical functions that link all model outputs used in the calibration process to all parameters requiring estimation. In enforcing history-matching constraints during the calibration and post-calibration uncertainty analysis processes, the proxy model is run for the purposes of populating the Jacobian matrix, while the original model is run when testing parameter upgrades; the latter process is readily parallelized. Use of a proxy model in this fashion dramatically reduces the computational burden of complex model calibration and uncertainty analysis. At the same time, the effect of model numerical misbehaviour on calculation of local gradients is mitigated, this allowing access to the benefits of gradient-based analysis where lack of integrity in finite-difference derivatives calculation would otherwise have impeded such access. Construction of a proxy model, and its subsequent use in calibration of a complex model, and in analysing the uncertainties of predictions made by that model, is implemented in the PEST suite.

  10. Calibration Lessons Learned from Hyperion Experience

    Science.gov (United States)

    Casement, S.; Ho, K.; Sandor-Leahy, S.; Biggar, S.; Czapla-Myers, J.; McCorkel, J.; Thome, K.

    2009-12-01

    improved methodologies during sensor development and additional capabilities of for on-orbit calibration systems that will enable reduction of the error terms. A key lesson learned was the approach taken to combine the various calibration results of Hyperion which differed by more than uncertainties in the methods.

  11. New calibration methodology for calorimetric determination of isobaric thermal expansivity of liquids as a function of temperature and pressure

    Energy Technology Data Exchange (ETDEWEB)

    Navia, Paloma; Troncoso, Jacobo [Departamento de Fisica Aplicada, Facultad de Ciencias de Ourense, Campus As Lagoas, 32004 Ourense (Spain); Romani, Luis [Departamento de Fisica Aplicada, Facultad de Ciencias de Ourense, Campus As Lagoas, 32004 Ourense (Spain)], E-mail: romani@uvigo.es

    2008-11-15

    A new method for determining isobaric thermal expansivity of liquids as a function of temperature and pressure through calorimetric measurements against pressure is described. It is based on a previously reported measurement technique, but due to the different kind of calorimeter and experimental set up, a new calibration procedure was developed. Two isobaric thermal expansivity standards are needed; in this work, with a view on the quality of the available literature data, hexane and water are chosen. The measurements were carried out in the temperature and pressure intervals (278.15 to 348.15) K and (0.5 to 55) MPa for a set of liquids, and experimental values are compared with the available literature data in order to evaluate the precision of the experimental procedure. The analysis of the results reveals that the proposed methodology is highly accurate for isobaric thermal expansivity determination, and it allows obtaining a precise characterisation of the temperature and pressure dependence of this thermodynamic coefficient.

  12. New calibration methodology for calorimetric determination of isobaric thermal expansivity of liquids as a function of temperature and pressure

    International Nuclear Information System (INIS)

    Navia, Paloma; Troncoso, Jacobo; Romani, Luis

    2008-01-01

    A new method for determining isobaric thermal expansivity of liquids as a function of temperature and pressure through calorimetric measurements against pressure is described. It is based on a previously reported measurement technique, but due to the different kind of calorimeter and experimental set up, a new calibration procedure was developed. Two isobaric thermal expansivity standards are needed; in this work, with a view on the quality of the available literature data, hexane and water are chosen. The measurements were carried out in the temperature and pressure intervals (278.15 to 348.15) K and (0.5 to 55) MPa for a set of liquids, and experimental values are compared with the available literature data in order to evaluate the precision of the experimental procedure. The analysis of the results reveals that the proposed methodology is highly accurate for isobaric thermal expansivity determination, and it allows obtaining a precise characterisation of the temperature and pressure dependence of this thermodynamic coefficient

  13. Validation and calibration of structural models that combine information from multiple sources.

    Science.gov (United States)

    Dahabreh, Issa J; Wong, John B; Trikalinos, Thomas A

    2017-02-01

    Mathematical models that attempt to capture structural relationships between their components and combine information from multiple sources are increasingly used in medicine. Areas covered: We provide an overview of methods for model validation and calibration and survey studies comparing alternative approaches. Expert commentary: Model validation entails a confrontation of models with data, background knowledge, and other models, and can inform judgments about model credibility. Calibration involves selecting parameter values to improve the agreement of model outputs with data. When the goal of modeling is quantitative inference on the effects of interventions or forecasting, calibration can be viewed as estimation. This view clarifies issues related to parameter identifiability and facilitates formal model validation and the examination of consistency among different sources of information. In contrast, when the goal of modeling is the generation of qualitative insights about the modeled phenomenon, calibration is a rather informal process for selecting inputs that result in model behavior that roughly reproduces select aspects of the modeled phenomenon and cannot be equated to an estimation procedure. Current empirical research on validation and calibration methods consists primarily of methodological appraisals or case-studies of alternative techniques and cannot address the numerous complex and multifaceted methodological decisions that modelers must make. Further research is needed on different approaches for developing and validating complex models that combine evidence from multiple sources.

  14. Development of methodology for evaluation of 99mTc and 131I incorporated activities during lactation

    International Nuclear Information System (INIS)

    Santos, L.; Dantas, A.L.A.; Mesquita, S.A.; Oliveira, S.M.V.; Instituto de Radioprotecao e Dosimetria

    2012-01-01

    Internal contamination of babies may occur for milk ingestion or inhalation of mothers occupationally exposed to ionizing radiation with possible incorporation or mothers submitted to medical exposures during lactation. Radionuclide concentrations in the mother's milk may cause organ absorbed doses in the babies proportionally to the breast volumes. Milk analysis allow to determine activities ingested by the babies by determining the peak of mother's milk considering the decrease of the activity rate and the milk activities drunk at different time intervals. The work had the aim to develop simulators and methodology to evaluate 99m Tc and 131 I in lactation, in the following steps: to prepare standard solution of contaminated milk separately with 99m Tc and 131 I; to build four breast simulators (600 g and 800 g) and respective calibration for two geometries (breast and whole-body) in the Whole-Body Counter Unit in Instituto de Radioprotecao e Dosimetria. The results demonstrated the system efficiency to determine 99m Tc and 131 I activities in breasts during the lactation period. The methodology for positioning in the 'breast geometry' seemed to be more efficient than the 'whole-body geometry' for different breast volumes. The experiment allows achieving better evaluation of internal dosimetry of mothers and their young children. (author)

  15. Capital Structure Arbitrage under a Risk-Neutral Calibration

    Directory of Open Access Journals (Sweden)

    Peter J. Zeitsch

    2017-01-01

    Full Text Available By reinterpreting the calibration of structural models, a reassessment of the importance of the input variables is undertaken. The analysis shows that volatility is the key parameter to any calibration exercise, by several orders of magnitude. To maximize the sensitivity to volatility, a simple formulation of Merton’s model is proposed that employs deep out-of-the-money option implied volatilities. The methodology also eliminates the use of historic data to specify the default barrier, thereby leading to a full risk-neutral calibration. Subsequently, a new technique for identifying and hedging capital structure arbitrage opportunities is illustrated. The approach seeks to hedge the volatility risk, or vega, as opposed to the exposure from the underlying equity itself, or delta. The results question the efficacy of the common arbitrage strategy of only executing the delta hedge.

  16. Automatic Calibration and Reconstruction for Active Vision Systems

    CERN Document Server

    Zhang, Beiwei

    2012-01-01

    In this book, the design of two new planar patterns for camera calibration of intrinsic parameters is addressed and a line-based method for distortion correction is suggested. The dynamic calibration of structured light systems, which consist of a camera and a projector is also treated. Also, the 3D Euclidean reconstruction by using the image-to-world transformation is investigated. Lastly, linear calibration algorithms for the catadioptric camera are considered, and the homographic matrix and fundamental matrix are extensively studied. In these methods, analytic solutions are provided for the computational efficiency and redundancy in the data can be easily incorporated to improve reliability of the estimations. This volume will therefore prove valuable and practical tool for researchers and practioners working in image processing and computer vision and related subjects.

  17. Calibration belt for quality-of-care assessment based on dichotomous outcomes.

    Directory of Open Access Journals (Sweden)

    Stefano Finazzi

    Full Text Available Prognostic models applied in medicine must be validated on independent samples, before their use can be recommended. The assessment of calibration, i.e., the model's ability to provide reliable predictions, is crucial in external validation studies. Besides having several shortcomings, statistical techniques such as the computation of the standardized mortality ratio (SMR and its confidence intervals, the Hosmer-Lemeshow statistics, and the Cox calibration test, are all non-informative with respect to calibration across risk classes. Accordingly, calibration plots reporting expected versus observed outcomes across risk subsets have been used for many years. Erroneously, the points in the plot (frequently representing deciles of risk have been connected with lines, generating false calibration curves. Here we propose a methodology to create a confidence band for the calibration curve based on a function that relates expected to observed probabilities across classes of risk. The calibration belt allows the ranges of risk to be spotted where there is a significant deviation from the ideal calibration, and the direction of the deviation to be indicated. This method thus offers a more analytical view in the assessment of quality of care, compared to other approaches.

  18. LOX/hydrocarbon rocket engine analytical design methodology development and validation. Volume 2: Appendices

    Science.gov (United States)

    Niiya, Karen E.; Walker, Richard E.; Pieper, Jerry L.; Nguyen, Thong V.

    1993-05-01

    This final report includes a discussion of the work accomplished during the period from Dec. 1988 through Nov. 1991. The objective of the program was to assemble existing performance and combustion stability models into a usable design methodology capable of designing and analyzing high-performance and stable LOX/hydrocarbon booster engines. The methodology was then used to design a validation engine. The capabilities and validity of the methodology were demonstrated using this engine in an extensive hot fire test program. The engine used LOX/RP-1 propellants and was tested over a range of mixture ratios, chamber pressures, and acoustic damping device configurations. This volume contains time domain and frequency domain stability plots which indicate the pressure perturbation amplitudes and frequencies from approximately 30 tests of a 50K thrust rocket engine using LOX/RP-1 propellants over a range of chamber pressures from 240 to 1750 psia with mixture ratios of from 1.2 to 7.5. The data is from test configurations which used both bitune and monotune acoustic cavities and from tests with no acoustic cavities. The engine had a length of 14 inches and a contraction ratio of 2.0 using a 7.68 inch diameter injector. The data was taken from both stable and unstable tests. All combustion instabilities were spontaneous in the first tangential mode. Although stability bombs were used and generated overpressures of approximately 20 percent, no tests were driven unstable by the bombs. The stability instrumentation included six high-frequency Kistler transducers in the combustion chamber, a high-frequency Kistler transducer in each propellant manifold, and tri-axial accelerometers. Performance data is presented, both characteristic velocity efficiencies and energy release efficiencies, for those tests of sufficient duration to record steady state values.

  19. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities; Metodos de calibracao e de intercomparacao de calibradores de dose utilizados em servicos de medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Costa, Alessandro Martins da

    1999-07-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  20. Initial tank calibration at NUCEF critical facility. 1. Measurement procedure and its result

    International Nuclear Information System (INIS)

    Yanagisawa, Hiroshi; Mineo, Hideaki; Tonoike, Kotaro; Takeshita, Isao; Hoshi, Katsuya; Hagiwara, Hiroyuki.

    1994-07-01

    Initial tank calibrations were carried out prior to hot operation of critical facilities in NUCEF: Nuclear Fuel Cycle Safety Engineering Research Facility, for the purpose of the nuclear material accountancy and control for the facility. Raw calibration data were collected from single run per one tank by measuring differential pressure with dip-tube systems, weight of calibration liquid (demineralized water) poured into the tank, temperature in the tank and so on, without operation of tank ventilation system. Volume and level data were obtained by applying density and buoyancy corrections to the raw data. As a result, the evaluated measurement errors of volume and level were small enough, e.g. within 0.2 lit. and 1.0 mm, respectively, for Pu accountancy tanks. This paper summarizes the above-mentioned measurement procedures, collected data, data correction procedures and evaluated measurement errors. (author)

  1. Technical Note: New methodology for measuring viscosities in small volumes characteristic of environmental chamber particle samples

    Directory of Open Access Journals (Sweden)

    L. Renbaum-Wolff

    2013-01-01

    Full Text Available Herein, a method for the determination of viscosities of small sample volumes is introduced, with important implications for the viscosity determination of particle samples from environmental chambers (used to simulate atmospheric conditions. The amount of sample needed is < 1 μl, and the technique is capable of determining viscosities (η ranging between 10−3 and 103 Pascal seconds (Pa s in samples that cover a range of chemical properties and with real-time relative humidity and temperature control; hence, the technique should be well-suited for determining the viscosities, under atmospherically relevant conditions, of particles collected from environmental chambers. In this technique, supermicron particles are first deposited on an inert hydrophobic substrate. Then, insoluble beads (~1 μm in diameter are embedded in the particles. Next, a flow of gas is introduced over the particles, which generates a shear stress on the particle surfaces. The sample responds to this shear stress by generating internal circulations, which are quantified with an optical microscope by monitoring the movement of the beads. The rate of internal circulation is shown to be a function of particle viscosity but independent of the particle material for a wide range of organic and organic-water samples. A calibration curve is constructed from the experimental data that relates the rate of internal circulation to particle viscosity, and this calibration curve is successfully used to predict viscosities in multicomponent organic mixtures.

  2. Tourism Methodologies - New Perspectives, Practices and Procedures

    DEFF Research Database (Denmark)

    This volume offers methodological discussions within the multidisciplinary field of tourism and shows how tourism researchers develop and apply new tourism methodologies. The book is presented as an anthology, giving voice to many diverse researchers who reflect on tourism methodology in differen...... codings and analysis, and tapping into the global network of social media.......This volume offers methodological discussions within the multidisciplinary field of tourism and shows how tourism researchers develop and apply new tourism methodologies. The book is presented as an anthology, giving voice to many diverse researchers who reflect on tourism methodology in different...... in interview and field work situations, and how do we engage with the performative aspects of tourism as a field of study? The book acknowledges that research is also performance and that it constitutes an aspect of intervention in the situations and contexts it is trying to explore. This is an issue dealt...

  3. Experimental determination of the drywell volume: 1/5 scale pressure suppression test

    International Nuclear Information System (INIS)

    Pitts, J.H.

    1977-01-01

    The BWR Mk 1 1/5 scale drywell volume, including space in all ports up to the first flange, was experimentally determined to be 10.01 m 3 with an uncertainty of +- 0.03 m 3 . The method of measurement used a 0.040879 m 3 calibrated volume that was initially filled with helium to 36.5 MPa. The calibrated volume was then connected to the drywell and the pressures equalized. The volumes of the vent pipes, instrumentation ports, and either the steam inlet or nitrogen inlet were subtracted from the measured drywell volume to obtain the net active drywell volume. The net active drywell volume is 9.87 m 3 for air tests and 9.85 m 3 for steam tests

  4. Robust, 'blind', in-situ calibration of SODARs

    International Nuclear Information System (INIS)

    Bradley, S

    2008-01-01

    Doppler spectra for SODARs are an average weighted over a volume according to the generally unknown beam pattern. By tilting the SODAR at a number of known angles, the beam pattern 'kernel' can be obtained and hence an absolute (field) calibration. No assumptions are required and the method does not require any comparison against mast or other installations

  5. In-situ calibration of TFTR [Tokamak Fusion Test Reactor] neutron detectors

    International Nuclear Information System (INIS)

    Hendel, H.W.; Palladino, R.W.; Barnes, C.W.; Diesso, M.; Felt, J.S.; Jassby, D.L.; Johnson, L.C.; Ku, L.P.; Liu, Q.P.; Motley, R.W.; Murphy, H.B.; Murphy, J.; Nieschmidt, E.B.; Roberts, J.A.; Saito, T.; Strachan, J.D.; Waszazak, R.J.; Young, K.

    1990-03-01

    We report results of the TFTR fission detector calibration performed in December 1988. A NBS-traceable, remotely controlled 252 Cf neutron source was moved toroidally through the TFTR vacuum vessel. Detection efficiencies for two 235 U detectors were measured for 930 locations of the neutron point source in toroidal scans at 16 different major radii and vertical heights. These scans effectively simulated the volume-distributed plasma neutron source, and the volume-integrated detection efficiency was found to be insensitive to plasma position. The Campbell mode is useful due to its large overlap with the count rate mode and large dynamic range. The resulting absolute plasma neutron source calibration has an uncertainty of ± 13%. 21 refs., 23 figs., 4 tabs

  6. A photogrammetric methodology for estimating construction and demolition waste composition

    International Nuclear Information System (INIS)

    Heck, H.H.; Reinhart, D.R.; Townsend, T.; Seibert, S.; Medeiros, S.; Cochran, K.; Chakrabarti, S.

    2002-01-01

    Manual sorting of construction, demolition, and renovation (C and D) waste is difficult and costly. A photogrammetric method has been developed to analyze the composition of C and D waste that eliminates the need for physical contact with the waste. The only field data collected is the weight and volume of the solid waste in the storage container and a photograph of each side of the waste pile, after it is dumped on the tipping floor. The methodology was developed and calibrated based on manual sorting studies at three different landfills in Florida, where the contents of twenty roll-off containers filled with C and D waste were sorted. The component classifications used were wood, concrete, paper products, drywall, metals, insulation, roofing, plastic, flooring, municipal solid waste, land-clearing waste, and other waste. Photographs of each side of the waste pile were taken with a digital camera and the pictures were analyzed on a computer using Photoshop software. Photoshop was used to divide the picture into eighty cells composed of ten columns and eight rows. The component distribution of each cell was estimated and results were summed to get a component distribution for the pile. Two types of distribution factors were developed that allow the component volumes and weights to be estimated. One set of distribution factors was developed to correct the volume distributions and the second set was developed to correct the weight distributions. The bulk density of each of the waste components were determined and used to convert waste volumes to weights. (author)

  7. A photogrammetric methodology for estimating construction and demolition waste composition

    Energy Technology Data Exchange (ETDEWEB)

    Heck, H.H. [Florida Inst. of Technology, Dept. of divil Engineering, Melbourne, Florida (United States); Reinhart, D.R.; Townsend, T.; Seibert, S.; Medeiros, S.; Cochran, K.; Chakrabarti, S

    2002-06-15

    Manual sorting of construction, demolition, and renovation (C and D) waste is difficult and costly. A photogrammetric method has been developed to analyze the composition of C and D waste that eliminates the need for physical contact with the waste. The only field data collected is the weight and volume of the solid waste in the storage container and a photograph of each side of the waste pile, after it is dumped on the tipping floor. The methodology was developed and calibrated based on manual sorting studies at three different landfills in Florida, where the contents of twenty roll-off containers filled with C and D waste were sorted. The component classifications used were wood, concrete, paper products, drywall, metals, insulation, roofing, plastic, flooring, municipal solid waste, land-clearing waste, and other waste. Photographs of each side of the waste pile were taken with a digital camera and the pictures were analyzed on a computer using Photoshop software. Photoshop was used to divide the picture into eighty cells composed of ten columns and eight rows. The component distribution of each cell was estimated and results were summed to get a component distribution for the pile. Two types of distribution factors were developed that allow the component volumes and weights to be estimated. One set of distribution factors was developed to correct the volume distributions and the second set was developed to correct the weight distributions. The bulk density of each of the waste components were determined and used to convert waste volumes to weights. (author)

  8. Criteria for the development and use of the methodology for environmentally-acceptable fossil energy site evaluation and selection. Volume 2. Final report

    Energy Technology Data Exchange (ETDEWEB)

    Eckstein, L.; Northrop, G.; Scott, R.

    1980-02-01

    This report serves as a companion document to the report, Volume 1: Environmentally-Acceptable Fossil Energy Site Evaluation and Selection: Methodology and Users Guide, in which a methodology was developed which allows the siting of fossil fuel conversion facilities in areas with the least environmental impact. The methodology, known as SELECS (Site Evaluation for Energy Conversion Systems) does not replace a site specific environmental assessment, or an environmental impact statement (EIS), but does enhance the value of an EIS by thinning down the number of options to a manageable level, by doing this in an objective, open and selective manner, and by providing preliminary assessment and procedures which can be utilized during the research and writing of the actual impact statement.

  9. Cuadernos de Autoformacion en Participacion Social: Metodologia. Volumen 2. Primera Edicion (Self-Instructional Notebooks on Social Participation: Methodology. Volume 2. First Edition).

    Science.gov (United States)

    Instituto Nacional para la Educacion de los Adultos, Mexico City (Mexico).

    The series "Self-instructional Notes on Social Participation" is a six-volume series intended as teaching aids for adult educators. The theoretical, methodological, informative and practical elements of this series will assist professionals in their work and help them achieve greater success. The specific purpose of each notebook is…

  10. Challenges in X-band Weather Radar Data Calibration

    DEFF Research Database (Denmark)

    Thorndahl, Søren; Rasmussen, Michael R.

    2009-01-01

    Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis and real time control purposes. In these contexts, it is all-important that the radar data well calibrated and adjusted in order to obtain valid quantitative precipitation e...... estimates. This paper compares two calibration procedures for a small marine X-band radar by comparing radar data with rain gauge data. Validation shows a very good consensus with regards to precipitation volumes, but more diverse results on peak rain intensities.......Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis and real time control purposes. In these contexts, it is all-important that the radar data well calibrated and adjusted in order to obtain valid quantitative precipitation...

  11. The Wally plot approach to assess the calibration of clinical prediction models.

    Science.gov (United States)

    Blanche, Paul; Gerds, Thomas A; Ekstrøm, Claus T

    2017-12-06

    A prediction model is calibrated if, roughly, for any percentage x we can expect that x subjects out of 100 experience the event among all subjects that have a predicted risk of x%. Typically, the calibration assumption is assessed graphically but in practice it is often challenging to judge whether a "disappointing" calibration plot is the consequence of a departure from the calibration assumption, or alternatively just "bad luck" due to sampling variability. We propose a graphical approach which enables the visualization of how much a calibration plot agrees with the calibration assumption to address this issue. The approach is mainly based on the idea of generating new plots which mimic the available data under the calibration assumption. The method handles the common non-trivial situations in which the data contain censored observations and occurrences of competing events. This is done by building on ideas from constrained non-parametric maximum likelihood estimation methods. Two examples from large cohort data illustrate our proposal. The 'wally' R package is provided to make the methodology easily usable.

  12. Calibration methods for the Hargreaves-Samani equation

    Directory of Open Access Journals (Sweden)

    Lucas Borges Ferreira

    Full Text Available ABSTRACT The estimation of the reference evapotranspiration is an important factor for hydrological studies, design and management of irrigation systems, among others. The Penman Monteith equation presents high precision and accuracy in the estimation of this variable. However, its use becomes limited due to the large number of required meteorological data. In this context, the Hargreaves-Samani equation could be used as alternative, although, for a better performance a local calibration is required. Thus, the aim was to compare the calibration process of the Hargreaves-Samani equation by linear regression, by adjustment of the coefficients (A and B and exponent (C of the equation and by combinations of the two previous alternatives. Daily data from 6 weather stations, located in the state of Minas Gerais, from the period 1997 to 2016 were used. The calibration of the Hargreaves-Samani equation was performed in five ways: calibration by linear regression, adjustment of parameter “A”, adjustment of parameters “A” and “C”, adjustment of parameters “A”, “B” and “C” and adjustment of parameters “A”, “B” and “C” followed by calibration by linear regression. The performances of the models were evaluated based on the statistical indicators mean absolute error, mean bias error, Willmott’s index of agreement, correlation coefficient and performance index. All the studied methodologies promoted better estimations of reference evapotranspiration. The simultaneous adjustment of the empirical parameters “A”, “B” and “C” was the best alternative for calibration of the Hargreaves-Samani equation.

  13. Correction volumes and densities in Vitrea Program

    International Nuclear Information System (INIS)

    Abrantes, Marcos E.S.; Oliveira, A.H. de

    2014-01-01

    Introduction: with the increased use of 3D reconstruction techniques to assist in diagnosis, Vitrea® program is widely used. To use this program you need to know the correction values to generate the volumes and number of real CT human tissues. Objective: provide correction values for volumes and number of CT, read the Vitrea program, of the tissues generated by DICOM images from CT. Methodology: this study used a PMMA chest phantom to generate the DICOM images on a scanner. To check the calibration of the scanner was used Catphan phantom and compared the manufacturer of the values associated with its straight linearity. Results: the volume of PMMA phantom was of 11166.58 cm³ and CT number (123.5 ± 33.4) UH. For the volume found in Vitrea program, according to the structures of interest, were 11897.29 cm 3 , 10901.65 cm³, 16906.49 cm 3 and 11848.34 cm³ and corrections values are -6.14%, + 2.43% -6.94% -5.75% respectively for the tissues: lung, bone, soft and full. For the CT numbers found in this program were (97.60 ± 58.9) UH, (72.00 ± 176.00) UH, (143.20 ± 19.50) UH and (31.90 ± 239,10) UH and corrections of + 26.54%, + 71.53%, -13.64% and 387.15% respectively for tissues: lung, bone, soft and full. Conclusion: the procedure performed can be used in other 3D reconstruction programs and where there are tools to reading CT number, observing the necessary corrections

  14. Calibrating electromagnetic induction conductivities with time-domain reflectometry measurements

    Science.gov (United States)

    Dragonetti, Giovanna; Comegna, Alessandro; Ajeel, Ali; Piero Deidda, Gian; Lamaddalena, Nicola; Rodriguez, Giuseppe; Vignoli, Giulio; Coppola, Antonio

    2018-02-01

    This paper deals with the issue of monitoring the spatial distribution of bulk electrical conductivity, σb, in the soil root zone by using electromagnetic induction (EMI) sensors under different water and salinity conditions. To deduce the actual distribution of depth-specific σb from EMI apparent electrical conductivity (ECa) measurements, we inverted the data by using a regularized 1-D inversion procedure designed to manage nonlinear multiple EMI-depth responses. The inversion technique is based on the coupling of the damped Gauss-Newton method with truncated generalized singular value decomposition (TGSVD). The ill-posedness of the EMI data inversion is addressed by using a sharp stabilizer term in the objective function. This specific stabilizer promotes the reconstruction of blocky targets, thereby contributing to enhance the spatial resolution of the EMI results in the presence of sharp boundaries (otherwise smeared out after the application of more standard Occam-like regularization strategies searching for smooth solutions). Time-domain reflectometry (TDR) data are used as ground-truth data for calibration of the inversion results. An experimental field was divided into four transects 30 m long and 2.8 m wide, cultivated with green bean, and irrigated with water at two different salinity levels and using two different irrigation volumes. Clearly, this induces different salinity and water contents within the soil profiles. For each transect, 26 regularly spaced monitoring soundings (1 m apart) were selected for the collection of (i) Geonics EM-38 and (ii) Tektronix reflectometer data. Despite the original discrepancies in the EMI and TDR data, we found a significant correlation of the means and standard deviations of the two data series; in particular, after a low-pass spatial filtering of the TDR data. Based on these findings, this paper introduces a novel methodology to calibrate EMI-based electrical conductivities via TDR direct measurements. This

  15. High-frequency measurements of aeolian saltation flux: Field-based methodology and applications

    Science.gov (United States)

    Martin, Raleigh L.; Kok, Jasper F.; Hugenholtz, Chris H.; Barchyn, Thomas E.; Chamecki, Marcelo; Ellis, Jean T.

    2018-02-01

    Aeolian transport of sand and dust is driven by turbulent winds that fluctuate over a broad range of temporal and spatial scales. However, commonly used aeolian transport models do not explicitly account for such fluctuations, likely contributing to substantial discrepancies between models and measurements. Underlying this problem is the absence of accurate sand flux measurements at the short time scales at which wind speed fluctuates. Here, we draw on extensive field measurements of aeolian saltation to develop a methodology for generating high-frequency (up to 25 Hz) time series of total (vertically-integrated) saltation flux, namely by calibrating high-frequency (HF) particle counts to low-frequency (LF) flux measurements. The methodology follows four steps: (1) fit exponential curves to vertical profiles of saltation flux from LF saltation traps, (2) determine empirical calibration factors through comparison of LF exponential fits to HF number counts over concurrent time intervals, (3) apply these calibration factors to subsamples of the saltation count time series to obtain HF height-specific saltation fluxes, and (4) aggregate the calibrated HF height-specific saltation fluxes into estimates of total saltation fluxes. When coupled to high-frequency measurements of wind velocity, this methodology offers new opportunities for understanding how aeolian saltation dynamics respond to variability in driving winds over time scales from tens of milliseconds to days.

  16. Fermentation process tracking through enhanced spectral calibration modeling.

    Science.gov (United States)

    Triadaphillou, Sophia; Martin, Elaine; Montague, Gary; Norden, Alison; Jeffkins, Paul; Stimpson, Sarah

    2007-06-15

    The FDA process analytical technology (PAT) initiative will materialize in a significant increase in the number of installations of spectroscopic instrumentation. However, to attain the greatest benefit from the data generated, there is a need for calibration procedures that extract the maximum information content. For example, in fermentation processes, the interpretation of the resulting spectra is challenging as a consequence of the large number of wavelengths recorded, the underlying correlation structure that is evident between the wavelengths and the impact of the measurement environment. Approaches to the development of calibration models have been based on the application of partial least squares (PLS) either to the full spectral signature or to a subset of wavelengths. This paper presents a new approach to calibration modeling that combines a wavelength selection procedure, spectral window selection (SWS), where windows of wavelengths are automatically selected which are subsequently used as the basis of the calibration model. However, due to the non-uniqueness of the windows selected when the algorithm is executed repeatedly, multiple models are constructed and these are then combined using stacking thereby increasing the robustness of the final calibration model. The methodology is applied to data generated during the monitoring of broth concentrations in an industrial fermentation process from on-line near-infrared (NIR) and mid-infrared (MIR) spectrometers. It is shown that the proposed calibration modeling procedure outperforms traditional calibration procedures, as well as enabling the identification of the critical regions of the spectra with regard to the fermentation process.

  17. Methodology, Algorithms, and Emerging Tool for Automated Design of Intelligent Integrated Multi-Sensor Systems

    Directory of Open Access Journals (Sweden)

    Andreas König

    2009-11-01

    Full Text Available The emergence of novel sensing elements, computing nodes, wireless communication and integration technology provides unprecedented possibilities for the design and application of intelligent systems. Each new application system must be designed from scratch, employing sophisticated methods ranging from conventional signal processing to computational intelligence. Currently, a significant part of this overall algorithmic chain of the computational system model still has to be assembled manually by experienced designers in a time and labor consuming process. In this research work, this challenge is picked up and a methodology and algorithms for automated design of intelligent integrated and resource-aware multi-sensor systems employing multi-objective evolutionary computation are introduced. The proposed methodology tackles the challenge of rapid-prototyping of such systems under realization constraints and, additionally, includes features of system instance specific self-correction for sustained operation of a large volume and in a dynamically changing environment. The extension of these concepts to the reconfigurable hardware platform renders so called self-x sensor systems, which stands, e.g., for self-monitoring, -calibrating, -trimming, and -repairing/-healing systems. Selected experimental results prove the applicability and effectiveness of our proposed methodology and emerging tool. By our approach, competitive results were achieved with regard to classification accuracy, flexibility, and design speed under additional design constraints.

  18. Strategy for design NIR calibration sets based on process spectrum and model space: An innovative approach for process analytical technology.

    Science.gov (United States)

    Cárdenas, V; Cordobés, M; Blanco, M; Alcalà, M

    2015-10-10

    The pharmaceutical industry is under stringent regulations on quality control of their products because is critical for both, productive process and consumer safety. According to the framework of "process analytical technology" (PAT), a complete understanding of the process and a stepwise monitoring of manufacturing are required. Near infrared spectroscopy (NIRS) combined with chemometrics have lately performed efficient, useful and robust for pharmaceutical analysis. One crucial step in developing effective NIRS-based methodologies is selecting an appropriate calibration set to construct models affording accurate predictions. In this work, we developed calibration models for a pharmaceutical formulation during its three manufacturing stages: blending, compaction and coating. A novel methodology is proposed for selecting the calibration set -"process spectrum"-, into which physical changes in the samples at each stage are algebraically incorporated. Also, we established a "model space" defined by Hotelling's T(2) and Q-residuals statistics for outlier identification - inside/outside the defined space - in order to select objectively the factors to be used in calibration set construction. The results obtained confirm the efficacy of the proposed methodology for stepwise pharmaceutical quality control, and the relevance of the study as a guideline for the implementation of this easy and fast methodology in the pharma industry. Copyright © 2015 Elsevier B.V. All rights reserved.

  19. De Minimis waste impacts analysis methodology. IMPACTS - BRC user's guide and methodology for radioactive wastes below regulatory concern. Draft report for comment. Volume 2

    International Nuclear Information System (INIS)

    Forstom, J.M.; Goode, D.J.

    1986-07-01

    This report describes the methodology and computer program used by NRC to evaluate radiological impacts associated with petitions to have specific slightly contaminated radioactive waste streams designated as ''below regulatory concern.'' These wastes could be treated and disposed of at facilities which are not licensed for low-level radioactive waste management. The IMPACTS-BRC computer program is implemented on IBM-PC microcomputers using the FORTRAN programming language. Radiological impacts (doses) are estimated for several pathways including direct gamma radiation exposure, worker inhalation and exposure, offsite atmospheric and water releases, and intruder exposures. Annual impacts are calculated for the maximum individual, critical groups, and general population. The treatment and disposal options include onsite incineration, incineration at municipal and hazardous waste facilities, and disposal at sanitary landfills and hazardous waste landfills. Modifications to the program (from Volume 1) are primarily for microcomputer compatibility and to provide information needed to evaluate the petitions. Default environmental and facility parameters are developed representing conservative assumptions about site selection and operational procedures. In particular, the parameters of the groundwater pathway model are modified to represent more conservative assumptions than the original model (Volume 1)

  20. Optimization of SPECT calibration for quantification of images applied to dosimetry with iodine-131; Otimização da calibração em SPECT para a quantificação de imagens aplicada à dosimetria com iodo-131

    Energy Technology Data Exchange (ETDEWEB)

    Carvalho, Samira Marques de

    2018-04-01

    SPECT systems calibration plays an essential role in the accuracy of the quantification of images. In this work, in its first stage, an optimized SPECT calibration method was proposed for {sup 131}I studies, considering the partial volume effect (PVE) and the position of the calibration source. In the second stage, the study aimed to investigate the impact of count density and reconstruction parameters on the determination of the calibration factor and the quantification of the image in dosimetry studies, considering the reality of clinical practice in Brazil. In the final step, the study aimed evaluating the influence of several factors in the calibration for absorbed dose calculation using Monte Carlo simulations (MC) GATE code. Calibration was performed by determining a calibration curve (sensitivity versus volume) obtained by applying different thresholds. Then, the calibration factors were determined with an exponential function adjustment. Images were performed with high and low counts densities for several source positions within the simulator. To validate the calibration method, the calibration factors were used for absolute quantification of the total reference activities. The images were reconstructed adopting two approaches of different parameters, usually used in patient images. The methodology developed for the calibration of the tomographic system was easier and faster to implement than other procedures suggested to improve the accuracy of the results. The study also revealed the influence of the location of the calibration source, demonstrating better precision in the absolute quantification considering the location of the target region during the calibration of the system. The study applied in the Brazilian thyroid protocol suggests the revision of the calibration of the SPECT system, including different positions for the reference source, besides acquisitions considering the Signal to Noise Ratio (SNR) of the images. Finally, the doses obtained

  1. Least square methods and covariance matrix applied to the relative efficiency calibration of a Ge(Li) detector

    International Nuclear Information System (INIS)

    Geraldo, L.P.; Smith, D.L.

    1989-01-01

    The methodology of covariance matrix and square methods have been applied in the relative efficiency calibration for a Ge(Li) detector apllied in the relative efficiency calibration for a Ge(Li) detector. Procedures employed to generate, manipulate and test covariance matrices which serve to properly represent uncertainties of experimental data are discussed. Calibration data fitting using least square methods has been performed for a particular experimental data set. (author) [pt

  2. Calibration of a stochastic health evolution model using NHIS data

    Science.gov (United States)

    Gupta, Aparna; Li, Zhisheng

    2011-10-01

    This paper presents and calibrates an individual's stochastic health evolution model. In this health evolution model, the uncertainty of health incidents is described by a stochastic process with a finite number of possible outcomes. We construct a comprehensive health status index (HSI) to describe an individual's health status, as well as a health risk factor system (RFS) to classify individuals into different risk groups. Based on the maximum likelihood estimation (MLE) method and the method of nonlinear least squares fitting, model calibration is formulated in terms of two mixed-integer nonlinear optimization problems. Using the National Health Interview Survey (NHIS) data, the model is calibrated for specific risk groups. Longitudinal data from the Health and Retirement Study (HRS) is used to validate the calibrated model, which displays good validation properties. The end goal of this paper is to provide a model and methodology, whose output can serve as a crucial component of decision support for strategic planning of health related financing and risk management.

  3. Study of the isolability of leaks in a network depending on calibration of demands

    OpenAIRE

    Pérez Magrané, Ramon; Nejjari Akhi-Elarab, Fatiha; Puig Cayuela, Vicenç; Quevedo Casín, Joseba Jokin; Cugueró Escofet, Miquel Àngel; Cugueró Escofet, Miquel Àngel; Peralta, Antoni

    2011-01-01

    Using the simulation models a methodology was developed for leakage detection and localisation in a water distribution network [4]. Such methodology is based on fault detection and isolation techniques. Sensitivity matrix of pressures to leaks [8] is generated using simulation models [1]. Such models are available in water companies. The methodology showed promising results but two main issues appeared. Calibration in models and precision in sensors was required. Investment in sensors could b...

  4. SINGLE IMAGE CAMERA CALIBRATION IN CLOSE RANGE PHOTOGRAMMETRY FOR SOLDER JOINT ANALYSIS

    Directory of Open Access Journals (Sweden)

    D. Heinemann

    2016-06-01

    Full Text Available Printed Circuit Boards (PCB play an important role in the manufacturing of electronic devices. To ensure a correct function of the PCBs a certain amount of solder paste is needed during the placement of components. The aim of the current research is to develop an real-time, closed-loop solution for the analysis of the printing process where solder is printed onto PCBs. Close range photogrammetry allows for determination of the solder volume and a subsequent correction if necessary. Photogrammetry is an image based method for three dimensional reconstruction from two dimensional image data of an object. A precise camera calibration is indispensable for an accurate reconstruction. In our certain application it is not possible to use calibration methods with two dimensional calibration targets. Therefore a special calibration target was developed and manufactured, which allows for single image camera calibration.

  5. Calibration of hydrological models using flow-duration curves

    Directory of Open Access Journals (Sweden)

    I. K. Westerberg

    2011-07-01

    Full Text Available The degree of belief we have in predictions from hydrologic models will normally depend on how well they can reproduce observations. Calibrations with traditional performance measures, such as the Nash-Sutcliffe model efficiency, are challenged by problems including: (1 uncertain discharge data, (2 variable sensitivity of different performance measures to different flow magnitudes, (3 influence of unknown input/output errors and (4 inability to evaluate model performance when observation time periods for discharge and model input data do not overlap. This paper explores a calibration method using flow-duration curves (FDCs to address these problems. The method focuses on reproducing the observed discharge frequency distribution rather than the exact hydrograph. It consists of applying limits of acceptability for selected evaluation points (EPs on the observed uncertain FDC in the extended GLUE approach. Two ways of selecting the EPs were tested – based on equal intervals of discharge and of volume of water. The method was tested and compared to a calibration using the traditional model efficiency for the daily four-parameter WASMOD model in the Paso La Ceiba catchment in Honduras and for Dynamic TOPMODEL evaluated at an hourly time scale for the Brue catchment in Great Britain. The volume method of selecting EPs gave the best results in both catchments with better calibrated slow flow, recession and evaporation than the other criteria. Observed and simulated time series of uncertain discharges agreed better for this method both in calibration and prediction in both catchments. An advantage with the method is that the rejection criterion is based on an estimation of the uncertainty in discharge data and that the EPs of the FDC can be chosen to reflect the aims of the modelling application, e.g. using more/less EPs at high/low flows. While the method appears less sensitive to epistemic input/output errors than previous use of limits of

  6. Impact of SPECT corrections on 3D-dosimetry for liver transarterial radioembolization using the patient relative calibration methodology

    Energy Technology Data Exchange (ETDEWEB)

    Pacilio, Massimiliano, E-mail: mpacilio@scamilloforlanini.rm.it; Basile, Chiara [Department of Medical Physics, Azienda Ospedaliera San Camillo Forlanini, Rome 00152 (Italy); Ferrari, Mahila; Botta, Francesca; Cremonesi, Marta [Department of Medical Physics, Istituto Europeo di Oncologia, Milan 20141 (Italy); Chiesa, Carlo [Department of Nuclear Medicine, Istituto Nazionale Tumori IRCCS Foundation, Milan 20133 (Italy); Lorenzon, Leda; Becci, Domenico [Postgraduate School of Medical Physics, “Sapienza” University of Rome, Rome 00185 (Italy); Mira, Marta [Post graduate Health Physics School, University of Milan, Milan 20122 (Italy); Torres, Leonel Alberto; Vergara Gil, Alex [Department of Nuclear Medicine, Clinical Research Division of the Center of Isotopes (DIC-CENTIS), Havana 11100 (Cuba); Coca Perez, Marco [Department of PET-CT and Nuclear Medicine, Imaging Center Medscan-Concepciòn, Concepciòn 4070061 (Chile); Ljungberg, Michael [Department of Medical Radiation Physics, University of Lund, Lund 22100 (Sweden); Pani, Roberto [Department of Medico-surgical Sciences and Biotecnologies, “Sapienza” University of Rome, Rome 00185 (Italy)

    2016-07-15

    Purpose: Many centers aim to plan liver transarterial radioembolization (TARE) with dosimetry, even without CT-based attenuation correction (AC), or with unoptimized scatter correction (SC) methods. This work investigates the impact of presence vs absence of such corrections, and limited spatial resolution, on 3D dosimetry for TARE. Methods: Three voxelized phantoms were derived from CT images of real patients with different body sizes. Simulations of {sup 99m}Tc-SPECT projections were performed with the SIMIND code, assuming three activity distributions in the liver: uniform, inside a “liver’s segment,” or distributing multiple uptaking nodules (“nonuniform liver”), with a tumoral liver/healthy parenchyma ratio of 5:1. Projection data were reconstructed by a commercial workstation, with OSEM protocol not specifically optimized for dosimetry (spatial resolution of 12.6 mm), with/without SC (optimized, or with parameters predefined by the manufacturer; dual energy window), and with/without AC. Activity in voxels was calculated by a relative calibration, assuming identical microspheres and {sup 99m}Tc-SPECT counts spatial distribution. 3D dose distributions were calculated by convolution with {sup 90}Y voxel S-values, assuming permanent trapping of microspheres. Cumulative dose-volume histograms in lesions and healthy parenchyma from different reconstructions were compared with those obtained from the reference biodistribution (the “gold standard,” GS), assessing differences for D95%, D70%, and D50% (i.e., minimum value of the absorbed dose to a percentage of the irradiated volume). γ tool analysis with tolerance of 3%/13 mm was used to evaluate the agreement between GS and simulated cases. The influence of deep-breathing was studied, blurring the reference biodistributions with a 3D anisotropic gaussian kernel, and performing the simulations once again. Results: Differences of the dosimetric indicators were noticeable in some cases, always negative

  7. Modeling Approach for Estimating Co-Produced Water Volumes and Saltwater Disposal Volumes in Oklahoma

    Science.gov (United States)

    Murray, K. E.

    2016-12-01

    Management of produced fluids has become an important issue in Oklahoma because large volumes of saltwater are co-produced with oil and gas, and disposed into saltwater disposal wells at high rates. Petroleum production increased from 2009-2015, especially in central and north-central Oklahoma where the Mississippian and Hunton zones were redeveloped using horizontal wells and dewatering techniques that have led to a disproportional increase in produced water volumes. Improved management of co-produced water, including desalination for beneficial reuse and decreased saltwater disposal volumes, is only possible if spatial and temporal trends can be defined and related to the producing zones. It is challenging to quantify the volumes of co-produced water by region or production zone because co-produced water volumes are generally not reported. Therefore, the goal of this research is to estimate co-produced water volumes for 2008-present with an approach that can be replicated as petroleum production shifts to other regions. Oil and gas production rates from subsurface zones were multiplied by ratios of H2O:oil and H2O:gas for the respective zones. Initial H2O:oil and H2O:gas ratios were adjusted/calibrated, by zone, to maximize correlation of county-scale produced H2O estimates versus saltwater disposal volumes from 2013-2015. These calibrated ratios were then used to compute saltwater disposal volumes from 2008-2012 because of apparent data gaps in reported saltwater disposal volumes during that timeframe. This research can be used to identify regions that have the greatest need for produced water treatment systems. The next step in management of produced fluids is to explore optimal energy-efficient strategies that reduce deleterious effects.

  8. Effective radiation attenuation calibration for breast density: compression thickness influences and correction

    Directory of Open Access Journals (Sweden)

    Thomas Jerry A

    2010-11-01

    Full Text Available Abstract Background Calibrating mammograms to produce a standardized breast density measurement for breast cancer risk analysis requires an accurate spatial measure of the compressed breast thickness. Thickness inaccuracies due to the nominal system readout value and compression paddle orientation induce unacceptable errors in the calibration. Method A thickness correction was developed and evaluated using a fully specified two-component surrogate breast model. A previously developed calibration approach based on effective radiation attenuation coefficient measurements was used in the analysis. Water and oil were used to construct phantoms to replicate the deformable properties of the breast. Phantoms consisting of measured proportions of water and oil were used to estimate calibration errors without correction, evaluate the thickness correction, and investigate the reproducibility of the various calibration representations under compression thickness variations. Results The average thickness uncertainty due to compression paddle warp was characterized to within 0.5 mm. The relative calibration error was reduced to 7% from 48-68% with the correction. The normalized effective radiation attenuation coefficient (planar representation was reproducible under intra-sample compression thickness variations compared with calibrated volume measures. Conclusion Incorporating this thickness correction into the rigid breast tissue equivalent calibration method should improve the calibration accuracy of mammograms for risk assessments using the reproducible planar calibration measure.

  9. CONSTRUCTION OF A CALIBRATED PROBABILISTIC CLASSIFICATION CATALOG: APPLICATION TO 50k VARIABLE SOURCES IN THE ALL-SKY AUTOMATED SURVEY

    International Nuclear Information System (INIS)

    Richards, Joseph W.; Starr, Dan L.; Miller, Adam A.; Bloom, Joshua S.; Brink, Henrik; Crellin-Quick, Arien; Butler, Nathaniel R.

    2012-01-01

    With growing data volumes from synoptic surveys, astronomers necessarily must become more abstracted from the discovery and introspection processes. Given the scarcity of follow-up resources, there is a particularly sharp onus on the frameworks that replace these human roles to provide accurate and well-calibrated probabilistic classification catalogs. Such catalogs inform the subsequent follow-up, allowing consumers to optimize the selection of specific sources for further study and permitting rigorous treatment of classification purities and efficiencies for population studies. Here, we describe a process to produce a probabilistic classification catalog of variability with machine learning from a multi-epoch photometric survey. In addition to producing accurate classifications, we show how to estimate calibrated class probabilities and motivate the importance of probability calibration. We also introduce a methodology for feature-based anomaly detection, which allows discovery of objects in the survey that do not fit within the predefined class taxonomy. Finally, we apply these methods to sources observed by the All-Sky Automated Survey (ASAS), and release the Machine-learned ASAS Classification Catalog (MACC), a 28 class probabilistic classification catalog of 50,124 ASAS sources in the ASAS Catalog of Variable Stars. We estimate that MACC achieves a sub-20% classification error rate and demonstrate that the class posterior probabilities are reasonably calibrated. MACC classifications compare favorably to the classifications of several previous domain-specific ASAS papers and to the ASAS Catalog of Variable Stars, which had classified only 24% of those sources into one of 12 science classes.

  10. CONSTRUCTION OF A CALIBRATED PROBABILISTIC CLASSIFICATION CATALOG: APPLICATION TO 50k VARIABLE SOURCES IN THE ALL-SKY AUTOMATED SURVEY

    Energy Technology Data Exchange (ETDEWEB)

    Richards, Joseph W.; Starr, Dan L.; Miller, Adam A.; Bloom, Joshua S.; Brink, Henrik; Crellin-Quick, Arien [Astronomy Department, University of California, Berkeley, CA 94720-3411 (United States); Butler, Nathaniel R., E-mail: jwrichar@stat.berkeley.edu [School of Earth and Space Exploration, Arizona State University, Tempe, AZ 85287 (United States)

    2012-12-15

    With growing data volumes from synoptic surveys, astronomers necessarily must become more abstracted from the discovery and introspection processes. Given the scarcity of follow-up resources, there is a particularly sharp onus on the frameworks that replace these human roles to provide accurate and well-calibrated probabilistic classification catalogs. Such catalogs inform the subsequent follow-up, allowing consumers to optimize the selection of specific sources for further study and permitting rigorous treatment of classification purities and efficiencies for population studies. Here, we describe a process to produce a probabilistic classification catalog of variability with machine learning from a multi-epoch photometric survey. In addition to producing accurate classifications, we show how to estimate calibrated class probabilities and motivate the importance of probability calibration. We also introduce a methodology for feature-based anomaly detection, which allows discovery of objects in the survey that do not fit within the predefined class taxonomy. Finally, we apply these methods to sources observed by the All-Sky Automated Survey (ASAS), and release the Machine-learned ASAS Classification Catalog (MACC), a 28 class probabilistic classification catalog of 50,124 ASAS sources in the ASAS Catalog of Variable Stars. We estimate that MACC achieves a sub-20% classification error rate and demonstrate that the class posterior probabilities are reasonably calibrated. MACC classifications compare favorably to the classifications of several previous domain-specific ASAS papers and to the ASAS Catalog of Variable Stars, which had classified only 24% of those sources into one of 12 science classes.

  11. Sensitivity analysis and development of calibration methodology for near-surface hydrogeology model of Laxemar

    International Nuclear Information System (INIS)

    Aneljung, Maria; Sassner, Mona; Gustafsson, Lars-Goeran

    2007-11-01

    This report describes modelling where the hydrological modelling system MIKE SHE has been used to describe surface hydrology, near-surface hydrogeology, advective transport mechanisms, and the contact between groundwater and surface water within the SKB site investigation area at Laxemar. In the MIKE SHE system, surface water flow is described with the one-dimensional modelling tool MIKE 11, which is fully and dynamically integrated with the groundwater flow module in MIKE SHE. In early 2008, a supplementary data set will be available and a process of updating, rebuilding and calibrating the MIKE SHE model based on this data set will start. Before the calibration on the new data begins, it is important to gather as much knowledge as possible on calibration methods, and to identify critical calibration parameters and areas within the model that require special attention. In this project, the MIKE SHE model has been further developed. The model area has been extended, and the present model also includes an updated bedrock model and a more detailed description of the surface stream network. The numerical model has been updated and optimized, especially regarding the modelling of evapotranspiration and the unsaturated zone, and the coupling between the surface stream network in MIKE 11 and the overland flow in MIKE SHE. An initial calibration has been made and a base case has been defined and evaluated. In connection with the calibration, the most important changes made in the model were the following: The evapotranspiration was reduced. The infiltration capacity was reduced. The hydraulic conductivities of the Quaternary deposits in the water-saturated part of the subsurface were reduced. Data from one surface water level monitoring station, four surface water discharge monitoring stations and 43 groundwater level monitoring stations (SSM series boreholes) have been used to evaluate and calibrate the model. The base case simulations showed a reasonable agreement

  12. Cross calibration of the Landsat-7 ETM+ and EO-1 ALI sensor

    Science.gov (United States)

    Chander, G.; Meyer, D.J.; Helder, D.L.

    2004-01-01

    As part of the Earth Observer 1 (EO-1) Mission, the Advanced Land Imager (ALI) demonstrates a potential technological direction for Landsat Data Continuity Missions. To evaluate ALI's capabilities in this role, a cross-calibration methodology has been developed using image pairs from the Landsat-7 (L7) Enhanced Thematic Mapper Plus (ETM+) and EO-1 (ALI) to verify the radiometric calibration of ALI with respect to the well-calibrated L7 ETM+ sensor. Results have been obtained using two different approaches. The first approach involves calibration of nearly simultaneous surface observations based on image statistics from areas observed simultaneously by the two sensors. The second approach uses vicarious calibration techniques to compare the predicted top-of-atmosphere radiance derived from ground reference data collected during the overpass to the measured radiance obtained from the sensor. The results indicate that the relative sensor chip assemblies gains agree with the ETM+ visible and near-infrared bands to within 2% and the shortwave infrared bands to within 4%.

  13. Using inductively coupled plasma-mass spectrometry for calibration transfer between environmental CRMs.

    Science.gov (United States)

    Turk, G C; Yu, L L; Salit, M L; Guthrie, W F

    2001-06-01

    Multielement analyses of environmental reference materials have been performed using existing certified reference materials (CRMs) as calibration standards for inductively coupled plasma-mass spectrometry. The analyses have been performed using a high-performance methodology that results in comparison measurement uncertainties that are significantly less than the uncertainties of the certified values of the calibration CRM. Consequently, the determined values have uncertainties that are very nearly equivalent to the uncertainties of the calibration CRM. Several uses of this calibration transfer are proposed, including, re-certification measurements of replacement CRMs, establishing traceability of one CRM to another, and demonstrating the equivalence of two CRMs. RM 8704, a river sediment, was analyzed using SRM 2704, Buffalo River Sediment, as the calibration standard. SRM 1632c, Trace Elements in Bituminous Coal, which is a replacement for SRM 1632b, was analyzed using SRM 1632b as the standard. SRM 1635, Trace Elements in Subbituminous Coal, was also analyzed using SRM 1632b as the standard.

  14. New Submersed Chamber for Calibration of Relative Humidity Instruments at HMI/FSB-LPM

    Science.gov (United States)

    Sestan, D.; Zvizdic, D.; Sariri, K.

    2018-02-01

    This paper gives a detailed description of a new chamber designed for calibration of relative humidity (RH) instruments at Laboratory for Process Measurement (HMI/FSB-LPM). To the present time, the calibrations of RH instruments at the HMI/FSB-LPM were done by comparison method using a climatic chamber of large volume and calibrated dew point hygrometer with an additional thermometer. Since 2010, HMI/FSB-LPM in cooperation with Centre for Metrology and Accreditation in Finland (MIKES) developed the two primary dew point generators which cover the dew point temperature range between - 70 {°}C and 60 {°}C. In order to utilize these facilities for calibrations of the RH instruments, the new chamber was designed, manufactured and installed in the existing system, aiming to extend its range and reduce the related calibration uncertainties. The chamber construction allows its use in a thermostatic bath of larger volume as well as in the climatic chambers. In the scope of this paper, performances of the new chamber were tested while it was submersed in a thermostated bath. The chamber can simultaneously accommodate up to three RH sensors. In order to keep the design of the chamber simple, only cylindrical RH sensors detachable from display units can be calibrated. Possible optimizations are also discussed, and improvements in the design proposed. By using the new chamber, HMI/FSB-LPM reduced the expanded calibration uncertainties (level of confidence 95 %, coverage factor k=2) from 0.6 %rh to 0.25 %rh at 30 %rh (23 {°}C), and from 0.8 %rh to 0.53 %rh at 70 %rh (23 {°}C).

  15. Principles and applications of measurement and uncertainty analysis in research and calibration

    Energy Technology Data Exchange (ETDEWEB)

    Wells, C.V.

    1992-11-01

    Interest in Measurement Uncertainty Analysis has grown in the past several years as it has spread to new fields of application, and research and development of uncertainty methodologies have continued. This paper discusses the subject from the perspectives of both research and calibration environments. It presents a history of the development and an overview of the principles of uncertainty analysis embodied in the United States National Standard, ANSI/ASME PTC 19.1-1985, Measurement Uncertainty. Examples are presented in which uncertainty analysis was utilized or is needed to gain further knowledge of a particular measurement process and to characterize final results. Measurement uncertainty analysis provides a quantitative estimate of the interval about a measured value or an experiment result within which the true value of that quantity is expected to lie. Years ago, Harry Ku of the United States National Bureau of Standards stated that ``The informational content of the statement of uncertainty determines, to a large extent, the worth of the calibrated value.`` Today, that statement is just as true about calibration or research results as it was in 1968. Why is that true? What kind of information should we include in a statement of uncertainty accompanying a calibrated value? How and where do we get the information to include in an uncertainty statement? How should we interpret and use measurement uncertainty information? This discussion will provide answers to these and other questions about uncertainty in research and in calibration. The methodology to be described has been developed by national and international groups over the past nearly thirty years, and individuals were publishing information even earlier. Yet the work is largely unknown in many science and engineering arenas. I will illustrate various aspects of uncertainty analysis with some examples drawn from the radiometry measurement and calibration discipline from research activities.

  16. Principles and applications of measurement and uncertainty analysis in research and calibration

    Energy Technology Data Exchange (ETDEWEB)

    Wells, C.V.

    1992-11-01

    Interest in Measurement Uncertainty Analysis has grown in the past several years as it has spread to new fields of application, and research and development of uncertainty methodologies have continued. This paper discusses the subject from the perspectives of both research and calibration environments. It presents a history of the development and an overview of the principles of uncertainty analysis embodied in the United States National Standard, ANSI/ASME PTC 19.1-1985, Measurement Uncertainty. Examples are presented in which uncertainty analysis was utilized or is needed to gain further knowledge of a particular measurement process and to characterize final results. Measurement uncertainty analysis provides a quantitative estimate of the interval about a measured value or an experiment result within which the true value of that quantity is expected to lie. Years ago, Harry Ku of the United States National Bureau of Standards stated that The informational content of the statement of uncertainty determines, to a large extent, the worth of the calibrated value.'' Today, that statement is just as true about calibration or research results as it was in 1968. Why is that true What kind of information should we include in a statement of uncertainty accompanying a calibrated value How and where do we get the information to include in an uncertainty statement How should we interpret and use measurement uncertainty information This discussion will provide answers to these and other questions about uncertainty in research and in calibration. The methodology to be described has been developed by national and international groups over the past nearly thirty years, and individuals were publishing information even earlier. Yet the work is largely unknown in many science and engineering arenas. I will illustrate various aspects of uncertainty analysis with some examples drawn from the radiometry measurement and calibration discipline from research activities.

  17. Calibration

    International Nuclear Information System (INIS)

    Greacen, E.L.; Correll, R.L.; Cunningham, R.B.; Johns, G.G.; Nicolls, K.D.

    1981-01-01

    Procedures common to different methods of calibration of neutron moisture meters are outlined and laboratory and field calibration methods compared. Gross errors which arise from faulty calibration techniques are described. The count rate can be affected by the dry bulk density of the soil, the volumetric content of constitutional hydrogen and other chemical components of the soil and soil solution. Calibration is further complicated by the fact that the neutron meter responds more strongly to the soil properties close to the detector and source. The differences in slope of calibration curves for different soils can be as much as 40%

  18. Monitoring angiogenesis using a human compatible calibration for broadband near-infrared spectroscopy

    Science.gov (United States)

    Yang, Runze; Zhang, Qiong; Wu, Ying; Dunn, Jeff F.

    2013-01-01

    Angiogenesis is a hallmark of many conditions, including cancer, stroke, vascular disease, diabetes, and high-altitude exposure. We have previously shown that one can study angiogenesis in animal models by using total hemoglobin (tHb) as a marker of cerebral blood volume (CBV), measured using broadband near-infrared spectroscopy (bNIRS). However, the method was not suitable for patients as global anoxia was used for the calibration. Here we determine if angiogenesis could be detected using a calibration method that could be applied to patients. CBV, as a marker of angiogenesis, is quantified in a rat cortex before and after hypoxia acclimation. Rats are acclimated at 370-mmHg pressure for three weeks, while rats in the control group are housed under the same conditions, but under normal pressure. CBV increased in each animal in the acclimation group. The mean CBV (%volume/volume) is 3.49%±0.43% (mean±SD) before acclimation for the experimental group, and 4.76%±0.29% after acclimation. The CBV for the control group is 3.28%±0.75%, and 3.09%±0.48% for the two measurements. This demonstrates that angiogenesis can be monitored noninvasively over time using a bNIRS system with a calibration method that is compatible with human use and less stressful for studies using animals.

  19. Calibration procedure for thermoluminescent dosemeters in water absorbed doses for Iridium-192 high dose rate sources

    International Nuclear Information System (INIS)

    Reyes Cac, Franky Eduardo

    2004-10-01

    Thermoluminescent dosimeters are used in brachytherapy services quality assurance programs, with the aim of guaranteeing the correct radiation dose supplied to cancer patients, as well as with the purpose of evaluating new clinical procedures. This work describes a methodology for thermoluminescent dosimeters calibration in terms of absorbed dose to water for 192 Ir high dose rate sources. The reference dose used is measured with an ionization chamber previously calibrated for 192 Ir energy quality, applying the methodology proposed by Toelli. This methodology aims to standardizing the procedure, in a similar form to that used for external radiotherapy. The work evolves the adaptation of the TRS-277 Code of the International Atomic Energy Agency, for small and big cavities, through the introduction for non-uniform experimental factor, for the absorbed dose in the neighborhood of small brachytherapy sources. In order to simulate a water medium around the source during the experimental work, an acrylic phantom was used. It guarantees the reproducibility of the ionization chamber and the thermoluminescent dosimeter's location in relation to the radiation source. The values obtained with the ionization chamber and the thermoluminescent dosimeters, exposed to a 192 Ir high dose rate source, were compared and correction factors for different source-detector distances were determined for the thermoluminescent dosimeters. A numeric function was generated relating the correction factors and the source-detector distance. These correction factors are in fact the thermoluminescent dosimeter calibration factors for the 192 Ir source considered. As a possible application of this calibration methodology for thermoluminescent dosimeters, a practical range of source-detector distances is proposed for quality control of 192 Ir high dose rate sources. (author)

  20. Sensitivity analysis and development of calibration methodology for near-surface hydrogeology model of Laxemar

    Energy Technology Data Exchange (ETDEWEB)

    Aneljung, Maria; Sassner, Mona; Gustafsson, Lars-Goeran (DHI Sverige AB, Lilla Bommen 1, SE-411 04 Goeteborg (Sweden))

    2007-11-15

    This report describes modelling where the hydrological modelling system MIKE SHE has been used to describe surface hydrology, near-surface hydrogeology, advective transport mechanisms, and the contact between groundwater and surface water within the SKB site investigation area at Laxemar. In the MIKE SHE system, surface water flow is described with the one-dimensional modelling tool MIKE 11, which is fully and dynamically integrated with the groundwater flow module in MIKE SHE. In early 2008, a supplementary data set will be available and a process of updating, rebuilding and calibrating the MIKE SHE model based on this data set will start. Before the calibration on the new data begins, it is important to gather as much knowledge as possible on calibration methods, and to identify critical calibration parameters and areas within the model that require special attention. In this project, the MIKE SHE model has been further developed. The model area has been extended, and the present model also includes an updated bedrock model and a more detailed description of the surface stream network. The numerical model has been updated and optimized, especially regarding the modelling of evapotranspiration and the unsaturated zone, and the coupling between the surface stream network in MIKE 11 and the overland flow in MIKE SHE. An initial calibration has been made and a base case has been defined and evaluated. In connection with the calibration, the most important changes made in the model were the following: The evapotranspiration was reduced. The infiltration capacity was reduced. The hydraulic conductivities of the Quaternary deposits in the water-saturated part of the subsurface were reduced. Data from one surface water level monitoring station, four surface water discharge monitoring stations and 43 groundwater level monitoring stations (SSM series boreholes) have been used to evaluate and calibrate the model. The base case simulations showed a reasonable agreement

  1. Estimation of the limit of detection in semiconductor gas sensors through linearized calibration models.

    Science.gov (United States)

    Burgués, Javier; Jiménez-Soto, Juan Manuel; Marco, Santiago

    2018-07-12

    The limit of detection (LOD) is a key figure of merit in chemical sensing. However, the estimation of this figure of merit is hindered by the non-linear calibration curve characteristic of semiconductor gas sensor technologies such as, metal oxide (MOX), gasFETs or thermoelectric sensors. Additionally, chemical sensors suffer from cross-sensitivities and temporal stability problems. The application of the International Union of Pure and Applied Chemistry (IUPAC) recommendations for univariate LOD estimation in non-linear semiconductor gas sensors is not straightforward due to the strong statistical requirements of the IUPAC methodology (linearity, homoscedasticity, normality). Here, we propose a methodological approach to LOD estimation through linearized calibration models. As an example, the methodology is applied to the detection of low concentrations of carbon monoxide using MOX gas sensors in a scenario where the main source of error is the presence of uncontrolled levels of humidity. Copyright © 2018 Elsevier B.V. All rights reserved.

  2. Calibration of snow avalanche mathematical models using the data of real avalanches in the Ile (Zailiyskiy Alatau Range

    Directory of Open Access Journals (Sweden)

    V. P. Blagoveshchensky

    2017-01-01

    Full Text Available The calibration of the dry friction and turbulent friction coefficients is necessary for computer simulation of avalanches. The method of back calculation based on data on actual avalanches is used for this purpose. The article presents the results of the calibration of the Eglit’s and RAMMS models for Ile Alatau range condi‑ tions. The range is located in Kazakhstan. The data on six avalanches in the same avalanche site were used. Five avalanches were dry, and one avalanche was wet. Avalanches volume varied from 2000 to 12000  m3. Maximum speed avalanches were between 15 and 30  m/s, the flow height  – from 3 to 10  m. Series of back calculations with different values of the friction coefficients was made to obtain the calibrated coeffi‑ cients. The calibrated coefficients were chosen under condition of the best fit with real avalanches. The cal‑ ibrated coefficients were following. For the Eglit’s model for dry avalanches of the volume 2000–5000  m3 μ = 0.46÷0.48, k = 0.005–0.006, and the volume 8000–12000 m3 μ = 0.38÷0.42, k = 0.002÷0.003. For RAMMS model for dry avalanches of the volume of 2000–5000 m3 μ (dry friction coefficient = 0.35÷0.4, ξ (viscous friction coefficient = 1500÷2000 m/s2, and the volume 8,000–12,000 m3 μ = 0.3÷0.35, ξ = 2000÷3000 m/s2. For wet avalanches of the volume 12,000 m3 μ = 0.35, ξ = 1500 m/s2. The work on the calibration will be con‑ tinued to obtain the friction coefficients for the Eglit’s and RAMMS models. The additional data on real ava‑ lanches will be needed for this purpose.

  3. Evaluation Methodology. The Evaluation Exchange. Volume 11, Number 2, Summer 2005

    Science.gov (United States)

    Coffman, Julia, Ed.

    2005-01-01

    This is the third issue of "The Evaluation Exchange" devoted entirely to the theme of methodology, though every issue tries to identify new methodological choices, the instructive ways in which people have applied or combined different methods, and emerging methodological trends. For example, lately "theories of change" have gained almost…

  4. Suomi-NPP VIIRS Day-Night Band On-Orbit Calibration and Performance

    Science.gov (United States)

    Chen, Hongda; Xiong, Xiaoxiong; Sun, Chengbo; Chen, Xuexia; Chiang, Kwofu

    2017-01-01

    The Suomi national polar-orbiting partnership Visible Infrared Imaging Radiometer Suite (VIIRS) instrument has successfully operated since its launch in October 2011. The VIIRS day-night band (DNB) is a panchromatic channel covering wavelengths from 0.5 to 0.9 microns that is capable of observing Earth scenes during both daytime and nighttime at a spatial resolution of 750 m. To cover the large dynamic range, the DNB operates at low-, middle-, and high-gain stages, and it uses an on-board solar diffuser (SD) for its low-gain stage calibration. The SD observations also provide a means to compute the gain ratios of low-to-middle and middle-to-high gain stages. This paper describes the DNB on-orbit calibration methodology used by the VIIRS characterization support team in supporting the NASA Earth science community with consistent VIIRS sensor data records made available by the land science investigator-led processing systems. It provides an assessment and update of the DNB on-orbit performance, including the SD degradation in the DNB spectral range, detector gain and gain ratio trending, and stray-light contamination and its correction. Also presented in this paper are performance validations based on Earth scenes and lunar observations, and comparisons to the calibration methodology used by the operational interface data processing segment.

  5. Evaluation of the accuracy of volume determination on the orbit and eyeball using MRI

    International Nuclear Information System (INIS)

    Chau, Anson; Fung, Karl; Yap, Maurice

    2005-01-01

    Purpose: This study reports a calibration carried out on phantoms simulating the orbit and eyeball to evaluate the accuracy of volumetric determination using MRI DICOM (Digital Imaging and Communication in Medicine) images. Methods: Ten tailor-made spherical silicon balls ranging in size from 5 to 14 cm 3 and 10 silicon moulds of orbits from 10 human dried skulls were used to simulate the eyes and orbits, respectively. The CISS (Constructive interference in steady state, TR/TE = 16/8 ms) T2-weighted sequence was taken using a Siemens MRI scanner. The volume of each phantom was computed and compared with the known physical volumes. Results: The computed and physical volumes were highly correlated for both eyeball (r = 0.997) and the orbit (r = 0.992) phantoms. Coefficients of variation of the computed and physical volumes were low. Consequently, it is possible to apply a calibration value to the computed volume to derive the physical volume with a high level of confidence. Conclusion: We conclude that with appropriate calibration, it is viable to use MRI DICOM images to derive the volume of the eyeball and the orbit

  6. Evaluation of the accuracy of volume determination on the orbit and eyeball using MRI

    Energy Technology Data Exchange (ETDEWEB)

    Chau, Anson [Department of Optometry and Radiography, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong (China); Fung, Karl [Department of Optometry and Radiography, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong (China)]. E-mail: orkarl@polyu.edu.hk; Yap, Maurice [Department of Optometry and Radiography, Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong (China)

    2005-02-01

    Purpose: This study reports a calibration carried out on phantoms simulating the orbit and eyeball to evaluate the accuracy of volumetric determination using MRI DICOM (Digital Imaging and Communication in Medicine) images. Methods: Ten tailor-made spherical silicon balls ranging in size from 5 to 14 cm{sup 3} and 10 silicon moulds of orbits from 10 human dried skulls were used to simulate the eyes and orbits, respectively. The CISS (Constructive interference in steady state, TR/TE = 16/8 ms) T2-weighted sequence was taken using a Siemens MRI scanner. The volume of each phantom was computed and compared with the known physical volumes. Results: The computed and physical volumes were highly correlated for both eyeball (r = 0.997) and the orbit (r = 0.992) phantoms. Coefficients of variation of the computed and physical volumes were low. Consequently, it is possible to apply a calibration value to the computed volume to derive the physical volume with a high level of confidence. Conclusion: We conclude that with appropriate calibration, it is viable to use MRI DICOM images to derive the volume of the eyeball and the orbit.

  7. Comparison of different methods for liquid level adjustment in tank prover calibration

    International Nuclear Information System (INIS)

    Garcia, D A; Farias, E C; Gabriel, P C; Aquino, M H; Gomes, R S E; Aibe, V Y

    2015-01-01

    The adjustment of the liquid level during the calibration of tank provers with fixed volume is normally done by overfill but it can be done in different ways. In this article four level adjustment techniques are compared: plate, pipette, ruler and overfill adjustment. The adjustment methods using plate and pipette presented good agreement with the tank's nominal volume and lower uncertainty among the tested methods

  8. Tritium monitor calibration at Los Alamos National Laboratory

    International Nuclear Information System (INIS)

    Bjork, C.J.; Aikin, D.J.; Houlton, T.W.

    1997-08-01

    Tritium in air is monitored at Los Alamos National Laboratory (LANL) with air breathing instruments based on ionization chambers. Stack emissions are continuously monitored from sample tubes which each connect to a Tritium bubble which differentially collects HTO and HT. A set of glass vials of glycol capture the HTO. The HT is oxidized with a palladium catalyst and the resultant HTO is captured in a second set of vials of glycol. The glycol is counted with a liquid scintillation counter. All calibrations are performed with tritium containing gas. The Radiation Instrumentation and Calibration (RIC) Team has constructed and maintains two closed loop gas handling systems based on femto TECH model U24 tritium ion chamber monitors: a fixed system housed in a fume hood and a portable system mounted on two two wheeled hand trucks. The U24 monitors are calibrated against tritium in nitrogen gas standards. They are used as standard transfer instruments to calibrate other ion chamber monitors with tritium in nitrogen, diluted with air. The gas handling systems include a circulation pump which permits a closed circulation loop to be established among the U24 monitor and typically two to four other monitors of a given model during calibration. Fixed and portable monitors can be calibrated. The stack bubblers are calibrated in the field by: blending a known concentration of tritium in air within the known volume of the two portable carts, coupled into a common loop; releasing that gas mixture into a ventilation intake to the stack; collecting oxidized tritium in the bubbler; counting the glycol; and using the stack and bubbler flow rates, computing the bubbler's efficiency. Gas calibration has become a convenient and quality tool in maintaining the tritium monitors at LANL

  9. Calibration method based on direct radioactivity measurement for radioactive gas monitoring instruments

    International Nuclear Information System (INIS)

    Yoshida, Makoto; Ohi, Yoshihiro; Chida, Tohru; Wu, Youyang.

    1993-01-01

    A calibration method for radioactive gas monitoring instruments was studied. In the method, gaseous radioactivity standards were provided on the basis of the direct radioactivity measurement by the diffusion-in long proportional counter method (DLPC method). The radioactivity concentration of the gas mixture through a monitoring instrument was determined by sampling the known volume of the gas mixture into the proportional counter used for the DLPC method. Since oxygen in the gas mixture decreased the counting efficiency in a proportional counter, the influence on calibration was experimentally estimated. It was not serious and able to be easily corrected. By the present method, the relation between radioactivity concentration and ionization current was determined for a gas-flow ionization chamber with 1.5 l effective volume. It showed good agreement with the results in other works. (author)

  10. A Study of IR Loss Correction Methodologies for Commercially Available Pyranometers

    Energy Technology Data Exchange (ETDEWEB)

    Long, Chuck; Andreas, Afshin; Augustine, John; Dooraghi, Mike; Habte, Aron; Hall, Emiel; Kutchenreiter, Mark; McComiskey, Allison; Reda, Ibrahim; Sengupta, Manajit

    2017-03-24

    This presentation provides a high-level overview of a study of IR Loss Connection Methodologies for Commercially Available Pyranometers. The IR Loss Corrections Study is investigating how various correction methodologies work for several makes and models of commercially available pyranometers in common use, both when operated in ventilators with DC fans and without ventilators, as when they are typically calibrated.

  11. Respiration monitoring by Electrical Bioimpedance (EBI) Technique in a group of healthy males. Calibration equations

    International Nuclear Information System (INIS)

    Balleza, M; Vargas, M; Delgadillo, I; Kashina, S; Huerta, M R; Moreno, G

    2017-01-01

    Several research groups have proposed the electrical impedance tomography (EIT) in order to analyse lung ventilation. With the use of 16 electrodes, the EIT is capable to obtain a set of transversal section images of thorax. In previous works, we have obtained an alternating signal in terms of impedance corresponding to respiration from EIT images. Then, in order to transform those impedance changes into a measurable volume signal a set of calibration equations has been obtained. However, EIT technique is still expensive to attend outpatients in basics hospitals. For that reason, we propose the use of electrical bioimpedance (EBI) technique to monitor respiration behaviour. The aim of this study was to obtain a set of calibration equations to transform EBI impedance changes determined at 4 different frequencies into a measurable volume signal. In this study a group of 8 healthy males was assessed. From obtained results, a high mathematical adjustment in the group calibrations equations was evidenced. Then, the volume determinations obtained by EBI were compared with those obtained by our gold standard. Therefore, despite EBI does not provide a complete information about impedance vectors of lung compared with EIT, it is possible to monitor the respiration. (paper)

  12. Research program for seismic qualification of nuclear plant electrical and mechanical equipment. Task 3. Recommendations for improvement of equipment qualification methodology and criteria. Volume 3

    International Nuclear Information System (INIS)

    Kana, D.D.; Pomerening, D.J.

    1984-08-01

    The Research Program for Seismic Qualification of Nuclear Plant Electrical and Mechanical Equipment has spanned a period of three years and resulted in seven technical summary reports, each of which covered in detail the findings of different tasks and subtasks, and have been combined into five NUREG/CR volumes. Volume 3 presents recommendations for improvement of equipment qualification methodology and procedural clarification/modification. The fifth category identifies issues where adequate information does not exist to allow a recommendation to be made

  13. Calibrating the radiation detector of the ventilation of a PET radiopharmaceutical production facility

    International Nuclear Information System (INIS)

    Lacerda, Marco Aurelio de Sousa; Tavares, Jose Carlos Freitas; Silva, Juliana Batista da

    2011-01-01

    The aim of this work is to demonstrate a new methodology of estimating the calibration factor of the ventilation duct of a PET radiopharmaceutical facility. The proposed methodology was studied to minimize contamination risks for the workers, as well as the uncertainties attributed to the gas sampling. The studied facility was the Development Centre of Nuclear Technology (CDTN/CNEN) in Belo Horizonte, Brazil. It was performed 3 consecutive irradiations with normal water (H 2 16 O) for production of nitrogen-13 to estimate the calibration factor of the detector located in the chimney of the facility. The readings of the detector were registered by the online radiation monitoring system (MEDISMARTS) during the transfer of the irradiated liquid until the count rate decreased for the background (BG) levels. The remaining activity of the water from the vial was measured and the decay corrected to the beginning of the transfer of the activity. The mean calibration factor estimated was (3.6 +- 0.5) kBq . m -3 . cps -1 . The maximum activities registered in the three irradiations were, respectively, 278 s, 370 s and 366 s after transferring of the activity to the hot cell. The conservative assumptions adopted and the values found for the calibration factor, which were close to the manufacturer published data, permit to estimate, safely, the discharges of radioactive gases in the installation. (author)

  14. Calibration of ARI QC ionisation chambers using the Australian secondary standards for activity

    International Nuclear Information System (INIS)

    Mo, L.; Van Der Gaast, H.A.; Alexiev, D.; Butcher, K.S.A.; Davies, J.

    1999-01-01

    The Secondary Standard Activity Laboratory (SSAL) in ANSTO routinely provides standardised radioactive sources, traceable activity measurements and custom source preparation services to customers. The most important activity carried out is the calibration of ionisation chambers located in the Quality Control (QC) section of Australian Radioisotopes (ARI). This ensures that their activity measurements are traceable to the Australian primary methods of standardisation. ARI QC ionisation chambers are calibrated for 99m Tc, 67 Ga, 131 I, 201 Tl and 153 Sm. The SSAL has a TPA ionisation chamber, which has been directly calibrated against a primary standard for a variety of radioactive nuclides. Calibration factors for this chamber were determined specifically for the actual volumes (5ml for 99m Tc, 131 I, 2ml for 67 Ga, 201 Tl and 3 ml for 153 Sm) and types of vial (Wheaton) which are routinely used at ARI. These calibration factors can be used to accurately measure the activity of samples prepared by ARI. The samples can subsequently be used to calibrate the QC ionisation chambers. QC ionisation chambers are re-calibrated biannually

  15. Development of a reference system and a methodology for the calibration of ophthalmic applicators utilized in brachytherapy

    International Nuclear Information System (INIS)

    Oliveira, Mercia Liane de

    2005-01-01

    90 Sr+ 90 Y beta radiation sources are widely utilized in brachytherapy, in the treatment of superficial lesions of eyes and skin. According to international recommendations, these applicators should be specified in terms of the absorbed dose rate to water at the reference point (1 mm from the source surface, along its axis of symmetry). Two mini-extrapolation chambers were developed with adequate geometrical characteristics for the dosimetry of plane and concave 90 Sr+ 90 Y sources. These chambers have 3.0 cm of outer diameter and 11.3 cm of length. Aluminized polyester foils are used as entrance windows, and the collecting electrodes were made of graphited polymethylmethacrylate. The mini-chambers were tested in 90 Sr+ 90 Y radiation beams from a beta check source and plane and concave ophthalmic applicators. All results obtained show the usefulness of these chambers as reference primary standards for the calibration of 90 Sr+ 90 Y applicators. The previous calibration of the mini-chambers in relation to a standard ionization chamber or to a standard beta source is unnecessary. The minichamber with plane window showed utility for low energy X-rays too. In order to establish an alternative method for the calibration of beta radiation sources, different thermoluminescent materials were tested: LiF, CaF 2 :Mn, CaF 2 :Dy and CaS0 4 :Dy. For their characterization, the response reproducibility, calibration curves, TL response as a function of the source-detector distance, transmission factors and the linearity of the sample response were determined. The calibration procedures of ophthalmic applicators were established utilizing the ionometric technique and thermoluminescence dosimetry. (author)

  16. External calibration of GOCE data using regional terrestrial gravity data

    Directory of Open Access Journals (Sweden)

    Wu Yunlong

    2012-08-01

    Full Text Available This paper reports on a study of the methodology of external calibration of GOCE data, using regional terrestrial-gravity data. Three regions around the world are selected in the numerical experiments. The result indicates that this calibration method is feasible. The effect is best with an accuracy of scale factor at 10−2 level, in Australia, where the area is smooth and the gravity data points are dense. The accuracy is one order of magnitude lower in both Canada where the area is smooth but the data points are sparse, and Norway, where the area is rather tough and the data points are sparse.

  17. Calibration of Correlation Radiometers Using Pseudo-Random Noise Signals

    Directory of Open Access Journals (Sweden)

    Sebastián Pantoja

    2009-08-01

    Full Text Available The calibration of correlation radiometers, and particularly aperture synthesis interferometric radiometers, is a critical issue to ensure their performance. Current calibration techniques are based on the measurement of the cross-correlation of receivers’ outputs when injecting noise from a common noise source requiring a very stable distribution network. For large interferometric radiometers this centralized noise injection approach is very complex from the point of view of mass, volume and phase/amplitude equalization. Distributed noise injection techniques have been proposed as a feasible alternative, but are unable to correct for the so-called “baseline errors” associated with the particular pair of receivers forming the baseline. In this work it is proposed the use of centralized Pseudo-Random Noise (PRN signals to calibrate correlation radiometers. PRNs are sequences of symbols with a long repetition period that have a flat spectrum over a bandwidth which is determined by the symbol rate. Since their spectrum resembles that of thermal noise, they can be used to calibrate correlation radiometers. At the same time, since these sequences are deterministic, new calibration schemes can be envisaged, such as the correlation of each receiver’s output with a baseband local replica of the PRN sequence, as well as new distribution schemes of calibration signals. This work analyzes the general requirements and performance of using PRN sequences for the calibration of microwave correlation radiometers, and particularizes the study to a potential implementation in a large aperture synthesis radiometer using an optical distribution network.

  18. New methodology for a person identification system

    Indian Academy of Sciences (India)

    Home; Journals; Sadhana; Volume 31; Issue 3. New methodology for a person identification system. R Bremananth A Chitra. Volume 31 Issue 3 June 2006 pp 259-276 ... Experimental results illustrate that the proposed method has been easily espoused in elections, bank transactions and other security applications.

  19. Calibration metrology for fixed irradiation sensors; Metrologie d'etalonnage des capteurs d'irradiation a poste fixe

    Energy Technology Data Exchange (ETDEWEB)

    Laloubere, L.; Ratsirahonana, A.; Djaroun, K. [CEA Saclay, 91 - Gif sur Yvette (France)

    2009-07-01

    After having recalled the regulatory and technical framework of the calibration of radioprotection measurement instruments, and outlined some technical and operational constraints, the authors report the development of an in situ calibration methodology, i.e. without displacement of the sensor. After the presentation of the calibration chain (from the measurement given by a fixed sensor to the reference value given by a primary standard), they indicate the definition and calculation of the different calibration coefficients allowing the linking up of the different levels, and also the taking of uncertainties into account. They finally report the validation of results

  20. Sensitivity analysis and development of calibration methodology for near-surface hydrogeology model of Forsmark

    International Nuclear Information System (INIS)

    Aneljung, Maria; Gustafsson, Lars-Goeran

    2007-04-01

    The hydrological modelling system MIKE SHE has been used to describe near-surface groundwater flow, transport mechanisms and the contact between ground- and surface water at the Forsmark site. The surface water system at Forsmark is described with the 1D modelling tool MIKE 11, which is fully and dynamically integrated with MIKE SHE. In spring 2007, a new data freeze will be available and a process of updating, rebuilding and calibrating the MIKE SHE model will start, based on the latest data set. Prior to this, it is important to gather as much knowledge as possible on calibration methods and to define critical calibration parameters and areas within the model. In this project, an optimization of the numerical description and an initial calibration of the MIKE SHE model has been made, and an updated base case has been defined. Data from 5 surface water level monitoring stations, 4 surface water discharge monitoring stations and 32 groundwater level monitoring stations (SFM soil boreholes) has been used for model calibration and evaluation. The base case simulations generally show a good agreement between calculated and measured water levels and discharges, indicating that the total runoff from the area is well described by the model. Moreover, with two exceptions (SFM0012 and SFM0022) the base case results show very good agreement between calculated and measured groundwater head elevations for boreholes installed below lakes. The model also shows a reasonably good agreement between calculated and measured groundwater head elevations or depths to phreatic surfaces in many other points. The following major types of calculation-measurement differences can be noted: Differences in groundwater level amplitudes due to transpiration processes. Differences in absolute mean groundwater head, due to differences between borehole casing levels and the interpolated DEM. Differences in absolute mean head elevations, due to local errors in hydraulic conductivity values

  1. Sensitivity analysis and development of calibration methodology for near-surface hydrogeology model of Forsmark

    Energy Technology Data Exchange (ETDEWEB)

    Aneljung, Maria; Gustafsson, Lars-Goeran [DHI Water and Environment AB, Goeteborg (Sweden)

    2007-04-15

    The hydrological modelling system MIKE SHE has been used to describe near-surface groundwater flow, transport mechanisms and the contact between ground- and surface water at the Forsmark site. The surface water system at Forsmark is described with the 1D modelling tool MIKE 11, which is fully and dynamically integrated with MIKE SHE. In spring 2007, a new data freeze will be available and a process of updating, rebuilding and calibrating the MIKE SHE model will start, based on the latest data set. Prior to this, it is important to gather as much knowledge as possible on calibration methods and to define critical calibration parameters and areas within the model. In this project, an optimization of the numerical description and an initial calibration of the MIKE SHE model has been made, and an updated base case has been defined. Data from 5 surface water level monitoring stations, 4 surface water discharge monitoring stations and 32 groundwater level monitoring stations (SFM soil boreholes) has been used for model calibration and evaluation. The base case simulations generally show a good agreement between calculated and measured water levels and discharges, indicating that the total runoff from the area is well described by the model. Moreover, with two exceptions (SFM0012 and SFM0022) the base case results show very good agreement between calculated and measured groundwater head elevations for boreholes installed below lakes. The model also shows a reasonably good agreement between calculated and measured groundwater head elevations or depths to phreatic surfaces in many other points. The following major types of calculation-measurement differences can be noted: Differences in groundwater level amplitudes due to transpiration processes. Differences in absolute mean groundwater head, due to differences between borehole casing levels and the interpolated DEM. Differences in absolute mean head elevations, due to local errors in hydraulic conductivity values

  2. Tornado missile simulation and design methodology. Volume 1: simulation methodology, design applications, and TORMIS computer code. Final report

    International Nuclear Information System (INIS)

    Twisdale, L.A.; Dunn, W.L.

    1981-08-01

    A probabilistic methodology has been developed to predict the probabilities of tornado-propelled missiles impacting and damaging nuclear power plant structures. Mathematical models of each event in the tornado missile hazard have been developed and sequenced to form an integrated, time-history simulation methodology. The models are data based where feasible. The data include documented records of tornado occurrence, field observations of missile transport, results of wind tunnel experiments, and missile impact tests. Probabilistic Monte Carlo techniques are used to estimate the risk probabilities. The methodology has been encoded in the TORMIS computer code to facilitate numerical analysis and plant-specific tornado missile probability assessments. Sensitivity analyses have been performed on both the individual models and the integrated methodology, and risk has been assessed for a hypothetical nuclear power plant design case study

  3. Camera Calibration of Stereo Photogrammetric System with One-Dimensional Optical Reference Bar

    International Nuclear Information System (INIS)

    Xu, Q Y; Ye, D; Che, R S; Qi, X; Huang, Y

    2006-01-01

    To carry out the precise measurement of large-scale complex workpieces, accurately calibration of the stereo photogrammetric system has becoming more and more important. This paper proposed a flexible and reliable camera calibration of stereo photogrammetric system based on quaternion with one-dimensional optical reference bar, which has three small collinear infrared LED marks and the lengths between these marks have been precisely calibration. By moving the optical reference bar at a number of locations/orientations over the measurement volume, we calibrate the stereo photogrammetric systems with the geometric constraint of the optical reference bar. The extrinsic parameters calibration process consists of linear parameters estimation based on quaternion and nonlinear refinement based on the maximum likelihood criterion. Firstly, we linear estimate the extrinsic parameters of the stereo photogrameetric systems based on quaternion. Then with the quaternion results as the initial values, we refine the extrinsic parameters through maximum likelihood criterion with the Levenberg-Marquardt Algorithm. In the calibration process, we can automatically control the light intensity and optimize the exposure time to get uniform intensity profile of the image points at different distance and obtain higher S/N ratio. The experiment result proves that the calibration method proposed is flexible, valid and obtains good results in the application

  4. NJP VOLUME 40 No 1B

    African Journals Online (AJOL)

    Prof Ezechukwu

    2012-07-13

    10.4314/njp.v40i1,9. Abstract Objective: Although the calibre of the airway is kept patent by multifactorial control system, there is evidence that ..... Lewinshohn HC, Capel LH, Smart. J. Changes in forced expiratory volumes throughout ...

  5. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 6: Accurate in-tank determination of liquid density in accountancy tanks equipped with dip tubes

    International Nuclear Information System (INIS)

    2008-01-01

    ISO 18213 deals with the acquisition, standardization, analysis, and use of calibration data to determine liquid volumes in process tanks for accountability purposes. This part of ISO 18213 is complementary to the other parts, ISO 18213-1 (procedural overview), ISO 18213-2 (data standardization), ISO 18213-3 (statistical methods), ISO 18213-4 (slow bubbling rate), ISO 18213-5 (fast bubbling rate). The procedure described in this part of ISO 18213 is a two-step procedure. First, a liquid of known density is used to determine the vertical distance between the tips of the two probes (i.e. to calibrate their separation). The calibration step requires synchronous (or as nearly synchronous as possible) measurements of the pressure exerted at the tips of two probes by the calibration liquid in which they are submerged. The measurements obtained are used to make an accurate determination of probe separation. Second, the unknown density of the process liquid is determined with the aid of the probe separation calibration. The density-determination step also requires (nearly) synchronous measurements of the pressure exerted at the tips of two probes by the process liquid of unknown density. With careful technique, it is possible to make determinations of liquid density with in-tank measurements that approach the accuracy and precision of those made in the laboratory. Moreover, density determinations made with in-tank measurements are automatically made at the observed temperature of the tank liquid. Thus, no additional information about the liquid is required to infer its density at its tank temperature from determinations of its density at some other temperature. Except that the density of the process liquid is generally not well characterized, the steps involved in determining the height of process liquid in the tank are the same as those for determining the height of calibration liquid. Thus, the method of density determination given in this part of ISO 18213 is very

  6. PIEZOELECTRIC WAVEGUIDE SENSOR FOR MEASURING PULSE PRESSURE IN CLOSED LIQUID VOLUMES AT HIGH VOLTAGE ELECTRIC DISCHARGE

    Directory of Open Access Journals (Sweden)

    V. G. Zhekul

    2017-10-01

    Full Text Available Purpose. Investigations of the characteristics of pressure waves presuppose the registration of the total profile of the pressure wave at a given point in space. For these purposes, various types of «pressure to the electrical signal» transmitters (sensors are used. Most of the common sensors are unsuitable for measuring the pulse pressure in a closed water volume at high hydrostatic pressures, in particular to study the effect of a powerful high-voltage pulse discharge on increasing the inflow of minerals and drinking water in wells. The purpose of the work was to develop antijamming piezoelectric waveguide sensor for measuring pulse pressure at a close distance from a high-voltage discharge channel in a closed volume of a liquid. Methodology. We have applied the calibration method as used as a secondary standard, the theory of electrical circuits. Results. We have selected the design and the circuit solution of the waveguide pressure sensor. We have developed a waveguide pulse-pressure sensor DTX-1 with a measuring loop. This sensor makes it possible to study the spectral characteristics of pressure waves of high-voltage pulse discharge in closed volumes of liquid at a hydrostatic pressure of up to 20 MPa and a temperature of up to 80 °C. The sensor can be used to study pressure waves with a maximum amplitude value of up to 150 MPa and duration of up to 80 µs. According to the results of the calibration, the sensitivity of the developed sensor DTX-1 with a measuring loop is 0.0346 V/MPa. Originality. We have further developed the theory of designing the waveguide piezoelectric pulse pressure sensors for measuring the pulse pressure at a close distance from a high-voltage discharge channel in a closed fluid volume by controlling the attenuation of the amplitude of the pressure signal. Practical value. We have developed, created, calibrated, used in scientific research waveguide pressure pulse sensors DTX-1. We propose sensors DTX-1 for sale

  7. Improved robotic stereotactic body radiation therapy plan quality and planning efficacy for organ-confined prostate cancer utilizing overlap-volume histogram-driven planning methodology

    International Nuclear Information System (INIS)

    Wu, Binbin; Pang, Dalong; Lei, Siyuan; Gatti, John; Tong, Michael; McNutt, Todd; Kole, Thomas; Dritschilo, Anatoly; Collins, Sean

    2014-01-01

    Background and purpose: This study is to determine if the overlap-volume histogram (OVH)-driven planning methodology can be adapted to robotic SBRT (CyberKnife Robotic Radiosurgery System) to further minimize the bladder and rectal doses achieved in plans manually-created by clinical planners. Methods and materials: A database containing clinically-delivered, robotic SBRT plans (7.25 Gy/fraction in 36.25 Gy) of 425 patients with localized prostate cancer was used as a cohort to establish an organ’s distance-to-dose model. The OVH-driven planning methodology was refined by adding the PTV volume factor to counter the target’s dose fall-off effect and incorporated into Multiplan to automate SBRT planning. For validation, automated plans (APs) for 12 new patients were generated, and their achieved dose/volume values were compared to the corresponding manually-created, clinically-delivered plans (CPs). A two-sided, Wilcoxon rank-sum test was used for statistical comparison with a significance level of p < 0.05. Results: PTV’s V(36.25 Gy) was comparable: 95.6% in CPs comparing to 95.1% in APs (p = 0.2). On average, the refined approach lowered V(18.12 Gy) to the bladder and rectum by 8.2% (p < 0.05) and 6.4% (p = 0.14). A physician confirmed APs were clinically acceptable. Conclusions: The improvements in APs could further reduce toxicities observed in SBRT for organ-confined prostate cancer

  8. Robust, 'blind', in-situ calibration of SODARs

    Energy Technology Data Exchange (ETDEWEB)

    Bradley, S [Physics Department, The University of Auckland, Private Bag 92019, Auckland (New Zealand); School of Computing, Science and Engineering, University of Salford, M5 4WT (United Kingdom)], E-mail: s.bradley@auckland.ac.nz

    2008-05-01

    Doppler spectra for SODARs are an average weighted over a volume according to the generally unknown beam pattern. By tilting the SODAR at a number of known angles, the beam pattern 'kernel' can be obtained and hence an absolute (field) calibration. No assumptions are required and the method does not require any comparison against mast or other installations.

  9. Probabilistic calibration of safety coefficients for flawed components in nuclear engineering

    International Nuclear Information System (INIS)

    Ardillon, E.; Pitner, P.; Barthelet, B.; Remond, A.

    1996-01-01

    The rules that are currently under application to verify the acceptance of flaws in nuclear components rely on deterministic criteria supposed to ensure the safe operating of plants. The interest of having a precise and reliable method to evaluate the safety margins and the integrity of components led Electricite de France to launch an approach to link directly safety coefficients with safety levels. This paper presents a probabilistic methodology to calibrate safety coefficients in relation to reliability target values. The proposed calibration procedure applies to the case of a ferritic flawed pipe using the R6 procedure for assessing the integrity of the structure. (authors). 5 refs., 5 figs

  10. Probabilistic calibration of safety coefficients for flawed components in nuclear engineering

    International Nuclear Information System (INIS)

    Ardillon, E.; Pitner, P.; Barthelet, B.; Remond, A.

    1995-01-01

    The current rules applied to verify the flaws acceptance in nuclear components rely on deterministic criteria supposed to ensure the plant safe operation. The interest in have a precise and reliable method to evaluate the safety margins and the integrity of components led Electricite de France to launch an approach to link directly safety coefficients with safety levels. This paper presents a probabilistic methodology to calibrate safety coefficients in relation do reliability target values. The proposed calibration procedure applies to the case of a ferritic flawed pipe using the R 6 procedure for assessing the structure integrity. (author). 5 refs., 5 figs., 1 tab

  11. The ethics of in vivo calibrations in oral health surveys.

    Science.gov (United States)

    Andrade, Flávia Reis de; Narvai, Paulo Capel; Montagner, Miguel Ângelo

    2016-01-01

    To analyze the ethics of in vivo calibration, using the discourse of the administrators of the National Oral Health Survey (SBBrasil 2010) as a starting point. This is a qualitative research involving semi-structured individual interviews with 12 members of the Steering Group and Technical Advisory Committee of the Ministry of Health, and two coordinators, one State and the other Municipal. The discourse of the collective subject technique was used for data analysis. When asked about the experiences of SBBrasil 2010, which included ethical aspects, respondents identified the forms of standardization and training of teams who collected field data. For them, there is little scientific evidence to ethically support the way the training stage, including calibration, is carried out in oral health epidemiological surveys, as a certain unease can be predicted in participants of these studies. The ethics of a research also derives from its methodological rigor; the training process; and calibration in particular, is a fundamental technical and ethical requirement in surveys such as the SBBrasil 2010. The unease of the volunteers in face of test repetition does not ethically invalidate the in vivo calibration, but mechanisms to minimize it must be developed.

  12. Calibration of a parsimonious distributed ecohydrological daily model in a data-scarce basin by exclusively using the spatio-temporal variation of NDVI

    Science.gov (United States)

    Ruiz-Pérez, Guiomar; Koch, Julian; Manfreda, Salvatore; Caylor, Kelly; Francés, Félix

    2017-12-01

    Ecohydrological modeling studies in developing countries, such as sub-Saharan Africa, often face the problem of extensive parametrical requirements and limited available data. Satellite remote sensing data may be able to fill this gap, but require novel methodologies to exploit their spatio-temporal information that could potentially be incorporated into model calibration and validation frameworks. The present study tackles this problem by suggesting an automatic calibration procedure, based on the empirical orthogonal function, for distributed ecohydrological daily models. The procedure is tested with the support of remote sensing data in a data-scarce environment - the upper Ewaso Ngiro river basin in Kenya. In the present application, the TETIS-VEG model is calibrated using only NDVI (Normalized Difference Vegetation Index) data derived from MODIS. The results demonstrate that (1) satellite data of vegetation dynamics can be used to calibrate and validate ecohydrological models in water-controlled and data-scarce regions, (2) the model calibrated using only satellite data is able to reproduce both the spatio-temporal vegetation dynamics and the observed discharge at the outlet and (3) the proposed automatic calibration methodology works satisfactorily and it allows for a straightforward incorporation of spatio-temporal data into the calibration and validation framework of a model.

  13. Metrological tests of a 200 L calibration source for HPGE detector systems for assay of radioactive waste drums

    International Nuclear Information System (INIS)

    Boshkova, T.; Mitev, K.

    2016-01-01

    In this work we present test procedures, approval criteria and results from two metrological inspections of a certified large volume "1"5"2Eu source (drum about 200 L) intended for calibration of HPGe gamma assay systems used for activity measurement of radioactive waste drums. The aim of the inspections was to prove the stability of the calibration source during its working life. The large volume source was designed and produced in 2007. It consists of 448 identical sealed radioactive sources (modules) apportioned in 32 transparent plastic tubes which were placed in a wooden matrix which filled the drum. During the inspections the modules were subjected to tests for verification of their certified characteristics. The results show a perfect compliance with the NIST basic guidelines for the properties of a radioactive certified reference material (CRM) and demonstrate the stability of the large volume CRM-drum after 7 years of operation. - Highlights: • Large (200 L) volume drum source designed, produced and certified as CRM in 2007. • Source contains 448 identical sealed radioactive "1"5"2Eu sources (modules). • Two metrological inspections in 2011 and 2014. • No statistically significant changes of the certified characteristics over time. • Stable calibration source for HPGe-gamma radioactive waste assay systems.

  14. SURF Model Calibration Strategy

    Energy Technology Data Exchange (ETDEWEB)

    Menikoff, Ralph [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-03-10

    SURF and SURFplus are high explosive reactive burn models for shock initiation and propagation of detonation waves. They are engineering models motivated by the ignition & growth concept of high spots and for SURFplus a second slow reaction for the energy release from carbon clustering. A key feature of the SURF model is that there is a partial decoupling between model parameters and detonation properties. This enables reduced sets of independent parameters to be calibrated sequentially for the initiation and propagation regimes. Here we focus on a methodology for tting the initiation parameters to Pop plot data based on 1-D simulations to compute a numerical Pop plot. In addition, the strategy for tting the remaining parameters for the propagation regime and failure diameter is discussed.

  15. Intercomparison of calibration procedures of high dose rate 192 Ir sources in Brazil and a proposal of a new methodology

    International Nuclear Information System (INIS)

    Marechal, M.H.; Almeida, C.E. de

    1998-01-01

    The objective of this paper is to report the results of an intercomparison of the calibration procedures for 192 Ir sources presently in use in Brazil and to proposal a calibration procedure to derive the N k for a Farmer type ionization chamber for 192 Ir energy by interpolating from a 60 Co gamma-rays and 250 kV x-rays calibration factors. the intercomparison results were all within ± 3.0 % except one case where 4.6 % was observed and latter identified as a problem with N-k value for X-rays. The method proposed by the present work make possible the improvement of the metrological coherence among the calibration laboratories and their users once the N k values could then provided by any of the members of SSDL network. (Author)

  16. A glimpse beneath Antarctic sea ice: observation of platelet-layer thickness and ice-volume fraction with multifrequency EM

    Science.gov (United States)

    Hoppmann, Mario; Hunkeler, Priska A.; Hendricks, Stefan; Kalscheuer, Thomas; Gerdes, Rüdiger

    2016-04-01

    In Antarctica, ice crystals (platelets) form and grow in supercooled waters below ice shelves. These platelets rise, accumulate beneath nearby sea ice, and subsequently form a several meter thick, porous sub-ice platelet layer. This special ice type is a unique habitat, influences sea-ice mass and energy balance, and its volume can be interpreted as an indicator of the health of an ice shelf. Although progress has been made in determining and understanding its spatio-temporal variability based on point measurements, an investigation of this phenomenon on a larger scale remains a challenge due to logistical constraints and a lack of suitable methodology. In the present study, we applied a lateral constrained Marquardt-Levenberg inversion to a unique multi-frequency electromagnetic (EM) induction sounding dataset obtained on the ice-shelf influenced fast-ice regime of Atka Bay, eastern Weddell Sea. We adapted the inversion algorithm to incorporate a sensor specific signal bias, and confirmed the reliability of the algorithm by performing a sensitivity study using synthetic data. We inverted the field data for sea-ice and platelet-layer thickness and electrical conductivity, and calculated ice-volume fractions within the platelet layer using Archie's Law. The thickness results agreed well with drillhole validation datasets within the uncertainty range, and the ice-volume fraction yielded results comparable to other studies. Both parameters together enable an estimation of the total ice volume within the platelet layer, which was found to be comparable to the volume of landfast sea ice in this region, and corresponded to more than a quarter of the annual basal melt volume of the nearby Ekström Ice Shelf. Our findings show that multi-frequency EM induction sounding is a suitable approach to efficiently map sea-ice and platelet-layer properties, with important implications for research into ocean/ice-shelf/sea-ice interactions. However, a successful application of this

  17. Design, manufacture, and calibration of infrared radiometric blackbody sources

    International Nuclear Information System (INIS)

    Byrd, D.A.; Michaud, F.D.; Bender, S.C.

    1996-04-01

    A Radiometric Calibration Station (RCS) is being assembled at the Los Alamos National Laboratories (LANL) which will allow for calibration of sensors with detector arrays having spectral capability from about 0.4-15 μm. The configuration of the LANL RCS. Two blackbody sources have been designed to cover the spectral range from about 3-15 μm, operating at temperatures ranging from about 180-350 K within a vacuum environment. The sources are designed to present a uniform spectral radiance over a large area to the sensor unit under test. The thermal uniformity requirement of the blackbody cavities has been one of the key factors of the design, requiring less than 50 mK variation over the entire blackbody surface to attain effective emissivity values of about 0.999. Once the two units are built and verified to the level of about 100 mK at LANL, they will be sent to the National Institute of Standards and Technology (NIST), where at least a factor of two improvement will be calibrated into the blackbody control system. The physical size of these assemblies will require modifications of the existing NIST Low Background Infrared (LBIR) Facility. LANL has constructed a bolt-on addition to the LBIR facility that will allow calibration of our large aperture sources. Methodology for attaining the two blackbody sources at calibrated levels of performance equivalent to present state of the art will be explained in the following

  18. UCalMiCeL – UNIFIED INTRINSIC AND EXTRINSIC CALIBRATION OF A MULTI-CAMERA-SYSTEM AND A LASERSCANNER

    Directory of Open Access Journals (Sweden)

    M. Hillemann

    2017-08-01

    Full Text Available Unmanned Aerial Vehicle (UAV with adequate sensors enable new applications in the scope between expensive, large-scale, aircraftcarried remote sensing and time-consuming, small-scale, terrestrial surveyings. To perform these applications, cameras and laserscanners are a good sensor combination, due to their complementary properties. To exploit this sensor combination the intrinsics and relative poses of the individual cameras and the relative poses of the cameras and the laserscanners have to be known. In this manuscript, we present a calibration methodology for the Unified Intrinsic and Extrinsic Calibration of a Multi-Camera-System and a Laserscanner (UCalMiCeL. The innovation of this methodology, which is an extension to the calibration of a single camera to a line laserscanner, is an unifying bundle adjustment step to ensure an optimal calibration of the entire sensor system. We use generic camera models, including pinhole, omnidirectional and fisheye cameras. For our approach, the laserscanner and each camera have to share a joint field of view, whereas the fields of view of the individual cameras may be disjoint. The calibration approach is tested with a sensor system consisting of two fisheye cameras and a line laserscanner with a range measuring accuracy of 30 mm. We evaluate the estimated relative poses between the cameras quantitatively by using an additional calibration approach for Multi-Camera-Systems based on control points which are accurately measured by a motion capture system. In the experiments, our novel calibration method achieves a relative pose estimation with a deviation below 1.8° and 6.4 mm.

  19. Metrology of the radon in air volume activity at the italian radon reference chamber

    Energy Technology Data Exchange (ETDEWEB)

    Sciocchetti, G.; Cotellessa, G.; Soldano, E.; Pagliari, M. [Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti, ENEA Centro Ricerche Casaccia Roma (Italy)

    2006-07-01

    The approach of the Italian National Institute of Ionising Radiations (I.N.M.R.I.-ENEA) on radon metrology has been based on a complete and integrated system which can be used to calibrate the main types of {sup 222}Rn in air measuring instruments with international traceability. The Italian radon reference chamber is a research and calibration facility developed at the Casaccia Research Center in Roma. This facility has an inner volume of one m{sup 3}. The wall is a cylindrical stainless steel vessel coupled with an automated climate apparatus operated both at steady and dynamic conditions. The control and data acquisition equipment is based on Radotron system, developed to automate the multitasking management of different sets of radon monitors and climatic sensors. A novel approach for testing passive radon monitors with an alpha track detector exposure standard has been developed. It is based on the direct measurement of radon exposure with a set of passive integrating monitors based on the new ENEA piston radon exposure meter. This paper describes the methodological approach on radon metrology, the status-of-art of experimental apparatus and the standardization procedures. (authors)

  20. Metrology of the radon in air volume activity at the italian radon reference chamber

    International Nuclear Information System (INIS)

    Sciocchetti, G.; Cotellessa, G.; Soldano, E.; Pagliari, M.

    2006-01-01

    The approach of the Italian National Institute of Ionising Radiations (I.N.M.R.I.-ENEA) on radon metrology has been based on a complete and integrated system which can be used to calibrate the main types of 222 Rn in air measuring instruments with international traceability. The Italian radon reference chamber is a research and calibration facility developed at the Casaccia Research Center in Roma. This facility has an inner volume of one m 3 . The wall is a cylindrical stainless steel vessel coupled with an automated climate apparatus operated both at steady and dynamic conditions. The control and data acquisition equipment is based on Radotron system, developed to automate the multitasking management of different sets of radon monitors and climatic sensors. A novel approach for testing passive radon monitors with an alpha track detector exposure standard has been developed. It is based on the direct measurement of radon exposure with a set of passive integrating monitors based on the new ENEA piston radon exposure meter. This paper describes the methodological approach on radon metrology, the status-of-art of experimental apparatus and the standardization procedures. (authors)

  1. Summary of KOMPSAT-5 Calibration and Validation

    Science.gov (United States)

    Yang, D.; Jeong, H.; Lee, S.; Kim, B.

    2013-12-01

    including pointing, relative and absolute calibration as well as geolocation accuracy determination. The absolute calibration will be accomplished by determining absolute radiometric accuracy using already deployed trihedral corner reflectors on calibration and validation sites located southeast from Ulaanbaatar, Mongolia. To establish a measure for the assess the final image products, geolocation accuracies of image products with different imaging modes will be determined by using deployed point targets and available Digital Terrain Model (DTM), and on different image processing levels. In summary, this paper will present calibration and validation activities performed during the LEOP and IOT of KOMPSAT-5. The methodology and procedure of calibration and validation will be explained as well as its results. Based on the results, the applications of SAR image products on geophysical processes will be also discussed.

  2. Drift-insensitive distributed calibration of probe microscope scanner in nanometer range: Virtual mode

    Science.gov (United States)

    Lapshin, Rostislav V.

    2016-08-01

    A method of distributed calibration of a probe microscope scanner is suggested. The main idea consists in a search for a net of local calibration coefficients (LCCs) in the process of automatic measurement of a standard surface, whereby each point of the movement space of the scanner can be characterized by a unique set of scale factors. Feature-oriented scanning (FOS) methodology is used as a basis for implementation of the distributed calibration permitting to exclude in situ the negative influence of thermal drift, creep and hysteresis on the obtained results. Possessing the calibration database enables correcting in one procedure all the spatial systematic distortions caused by nonlinearity, nonorthogonality and spurious crosstalk couplings of the microscope scanner piezomanipulators. To provide high precision of spatial measurements in nanometer range, the calibration is carried out using natural standards - constants of crystal lattice. One of the useful modes of the developed calibration method is a virtual mode. In the virtual mode, instead of measurement of a real surface of the standard, the calibration program makes a surface image ;measurement; of the standard, which was obtained earlier using conventional raster scanning. The application of the virtual mode permits simulation of the calibration process and detail analysis of raster distortions occurring in both conventional and counter surface scanning. Moreover, the mode allows to estimate the thermal drift and the creep velocities acting while surface scanning. Virtual calibration makes possible automatic characterization of a surface by the method of scanning probe microscopy (SPM).

  3. Innovative Calibration Method for System Level Simulation Models of Internal Combustion Engines

    Directory of Open Access Journals (Sweden)

    Ivo Prah

    2016-09-01

    Full Text Available The paper outlines a procedure for the computer-controlled calibration of the combined zero-dimensional (0D and one-dimensional (1D thermodynamic simulation model of a turbocharged internal combustion engine (ICE. The main purpose of the calibration is to determine input parameters of the simulation model in such a way as to achieve the smallest difference between the results of the measurements and the results of the numerical simulations with minimum consumption of the computing time. An innovative calibration methodology is based on a novel interaction between optimization methods and physically based methods of the selected ICE sub-systems. Therein physically based methods were used for steering the division of the integral ICE to several sub-models and for determining parameters of selected components considering their governing equations. Innovative multistage interaction between optimization methods and physically based methods allows, unlike the use of well-established methods that rely only on the optimization techniques, for successful calibration of a large number of input parameters with low time consumption. Therefore, the proposed method is suitable for efficient calibration of simulation models of advanced ICEs.

  4. Calibration of circular aperture area using vision probe at inmetro

    Directory of Open Access Journals (Sweden)

    Costa Pedro Bastos

    2016-01-01

    Full Text Available Circular aperture areas are standards of high importance for the realization of photometric and radiometric measurements, where the accuracy of these measures is related to the accuracy of the circular aperture area calibrations. In order to attend the requirement for traceability was developed in Brazilian metrology institute, a methodology for circular aperture area measurement as requirements from the radiometric and photometric measurements. In the developed methodology apertures are measured by non-contact measurement through images of the aperture edges captured by a camera. These images are processed using computer vision techniques and then the values of the circular aperture area are determined.

  5. Experimental procedures for the calibration of scintillation cells used in the determination of radon gas concentrations

    International Nuclear Information System (INIS)

    Grenier, M; Bigu, J.

    1982-02-01

    Experimental and analytical procedures are described for the calibration of scintillation cells used for the determination of radon gas concentration. In-house designed and built scintillation cells, used routinely in the monitoring of radon gas in uranium mine underground environments and in the laboratory, were calibrated. The cells had a volume of approximately 158 cm 3 and an α-counting efficiency ranging from 50% to 64%. Calibration factors for the cells were determined. Values ranged approximately from 0.177 cpm/pCiL -1 (4.77 cpm/BqL -1 ) to 0.224 cpm/pCiL -1 (6.05 cpm/BqL -1 ). The calibration facilities at the Elliot Lake Laboratory are briefly described

  6. Small-Volume Injections: Evaluation of Volume Administration Deviation From Intended Injection Volumes.

    Science.gov (United States)

    Muffly, Matthew K; Chen, Michael I; Claure, Rebecca E; Drover, David R; Efron, Bradley; Fitch, William L; Hammer, Gregory B

    2017-10-01

    In the perioperative period, anesthesiologists and postanesthesia care unit (PACU) nurses routinely prepare and administer small-volume IV injections, yet the accuracy of delivered medication volumes in this setting has not been described. In this ex vivo study, we sought to characterize the degree to which small-volume injections (≤0.5 mL) deviated from the intended injection volumes among a group of pediatric anesthesiologists and pediatric postanesthesia care unit (PACU) nurses. We hypothesized that as the intended injection volumes decreased, the deviation from those intended injection volumes would increase. Ten attending pediatric anesthesiologists and 10 pediatric PACU nurses each performed a series of 10 injections into a simulated patient IV setup. Practitioners used separate 1-mL tuberculin syringes with removable 18-gauge needles (Becton-Dickinson & Company, Franklin Lakes, NJ) to aspirate 5 different volumes (0.025, 0.05, 0.1, 0.25, and 0.5 mL) of 0.25 mM Lucifer Yellow (LY) fluorescent dye constituted in saline (Sigma Aldrich, St. Louis, MO) from a rubber-stoppered vial. Each participant then injected the specified volume of LY fluorescent dye via a 3-way stopcock into IV tubing with free-flowing 0.9% sodium chloride (10 mL/min). The injected volume of LY fluorescent dye and 0.9% sodium chloride then drained into a collection vial for laboratory analysis. Microplate fluorescence wavelength detection (Infinite M1000; Tecan, Mannedorf, Switzerland) was used to measure the fluorescence of the collected fluid. Administered injection volumes were calculated based on the fluorescence of the collected fluid using a calibration curve of known LY volumes and associated fluorescence.To determine whether deviation of the administered volumes from the intended injection volumes increased at lower injection volumes, we compared the proportional injection volume error (loge [administered volume/intended volume]) for each of the 5 injection volumes using a linear

  7. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration...

  8. Analysis and classification of data sets for calibration and validation of agro-ecosystem models

    DEFF Research Database (Denmark)

    Kersebaum, K C; Boote, K J; Jorgenson, J S

    2015-01-01

    Experimental field data are used at different levels of complexity to calibrate, validate and improve agro-ecosystem models to enhance their reliability for regional impact assessment. A methodological framework and software are presented to evaluate and classify data sets into four classes regar...

  9. Improved best estimate plus uncertainty methodology, including advanced validation concepts, to license evolving nuclear reactors

    International Nuclear Information System (INIS)

    Unal, C.; Williams, B.; Hemez, F.; Atamturktur, S.H.; McClure, P.

    2011-01-01

    Research highlights: → The best estimate plus uncertainty methodology (BEPU) is one option in the licensing of nuclear reactors. → The challenges for extending the BEPU method for fuel qualification for an advanced reactor fuel are primarily driven by schedule, the need for data, and the sufficiency of the data. → In this paper we develop an extended BEPU methodology that can potentially be used to address these new challenges in the design and licensing of advanced nuclear reactors. → The main components of the proposed methodology are verification, validation, calibration, and uncertainty quantification. → The methodology includes a formalism to quantify an adequate level of validation (predictive maturity) with respect to existing data, so that required new testing can be minimized, saving cost by demonstrating that further testing will not enhance the quality of the predictive tools. - Abstract: Many evolving nuclear energy technologies use advanced predictive multiscale, multiphysics modeling and simulation (M and S) capabilities to reduce the cost and schedule of design and licensing. Historically, the role of experiments has been as a primary tool for the design and understanding of nuclear system behavior, while M and S played the subordinate role of supporting experiments. In the new era of multiscale, multiphysics computational-based technology development, this role has been reversed. The experiments will still be needed, but they will be performed at different scales to calibrate and validate the models leading to predictive simulations for design and licensing. Minimizing the required number of validation experiments produces cost and time savings. The use of multiscale, multiphysics models introduces challenges in validating these predictive tools - traditional methodologies will have to be modified to address these challenges. This paper gives the basic aspects of a methodology that can potentially be used to address these new challenges in

  10. Study of calibration equations of 137Cs methodology for soil erosion determination

    International Nuclear Information System (INIS)

    Santos, Elias Antunes dos

    2001-02-01

    Using the method of 137 Cs and gamma-ray spectrometry, soil samples of two plots erosion were studied at Londrina city. the soil class studied was a dystrophic dark red soil (LRd), with erosion indexes measured by Agronomic Institute of Parana State (IAPAR) using a conventional method, since 1976. Through the percentage reduction of 137 Cs related to the reference site, the soil losses were calculated using the proportional, mass balance and profile distribution models. Making the correlation between the 137 Cs concentrations and the erosion measured by IAPAR, two calibration equations were obtained and applied to the data set measured in the basin of the Unda river and compared to those models in the literature. As reference region, was chosen a natural forest located close to the plots. The average inventory of 137 Cs was 555± 16 Bq.m -2 . The inventories of the erosion plots varied from 112 to 136 Bq.m -2 for samples collected until 30 cm depth. The erosion rates estimated by the models varied from 64 to 85 ton.ha -1 .yr -1 for the proportional and profile distribution models, respectively, and 137 to 165 ton.ha -1 for the mass balance model, while the measured erosion obtained by IAPAR was 86 ton.ha -1 .yr -1 . From the two calibration equations obtained, the one that take into account the 137 Cs distribution with the soil profile was that showed the best consistence with the erosion rated for the basin of the Unda river (same soil class) in the range from 4 to 48 ton.ha -1 .yr -1 , while the proportional and profile distribution models applied rates from 7 to 45 ton.ha -1 .yr -1 and 6 to 69 ton.ha -1 .yr -1 , respectively. (author)

  11. Measurement of tidal volume using respiratory ultrasonic plethysmography in anaesthetized, mechanically ventilated horses.

    Science.gov (United States)

    Russold, Elena; Ambrisko, Tamas D; Schramel, Johannes P; Auer, Ulrike; Van Den Hoven, Rene; Moens, Yves P

    2013-01-01

    To compare tidal volume estimations obtained from Respiratory Ultrasonic Plethysmography (RUP) with simultaneous spirometric measurements in anaesthetized, mechanically ventilated horses. Prospective randomized experimental study. Five experimental horses. Five horses were anaesthetized twice (1 week apart) in random order in lateral and in dorsal recumbency. Nine ventilation modes (treatments) were scheduled in random order (each lasting 4 minutes) applying combinations of different tidal volumes (8, 10, 12 mL kg(-1)) and positive end-expiratory pressures (PEEP) (0, 10, 20 cm H(2)O). Baseline ventilation mode (tidal volume=15 mL kg(-1), PEEP=0 cm H(2)O) was applied for 4 minutes between all treatments. Spirometry and RUP data were downloaded to personal computers. Linear regression analyses (RUP versus spirometric tidal volume) were performed using different subsets of data. Additonally RUP was calibrated against spirometry using a regression equation for all RUP signal values (thoracic, abdominal and combined) with all data collectively and also by an individually determined best regression equation (highest R(2)) for each experiment (horse versus recumbency) separately. Agreement between methods was assessed with Bland-Altman analyses. The highest correlation of RUP and spirometric tidal volume (R(2)=0.81) was found with the combined RUP signal in horses in lateral recumbency and ventilated without PEEP. The bias ±2 SD was 0±2.66 L when RUP was calibrated for collective data, but decreased to 0±0.87 L when RUP was calibrated with individual data. A possible use of RUP for tidal volume measurement during IPPV needs individual calibration to obtain limits of agreement within ±20%. © 2012 The Authors. Veterinary Anaesthesia and Analgesia. © 2012 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesiologists.

  12. Comparison of Tidal Volumes at the Endotracheal Tube and at the Ventilator.

    Science.gov (United States)

    Kim, Paul; Salazar, Adler; Ross, Patrick A; Newth, Christopher J L; Khemani, Robinder G

    2015-11-01

    Lung protective ventilation for children with acute respiratory distress syndrome requires accurate assessment of tidal volume. Although modern ventilators compensate for ventilator tubing compliance, tidal volume measured at the ventilator may not be accurate, particularly in small children. Although ventilator-specific proximal flow sensors that measure tidal volume at the endotracheal tube have been developed, there is little information regarding their accuracy. We sought to test the accuracy of ventilator measured tidal volume with and without proximal flow sensors against a calibrated pneumotachometer in children. Prospective, observational. Tertiary care PICU. Fifty-one endotracheally intubated and mechanically ventilated children younger than 18 years. Tidal volumes were measured at the ventilator, using a ventilator-specific flow sensor, and a calibrated pneumotachometer connected to the SensorMedics 2600A Pediatric Pulmonary Function Cart. In a pressure control mode of ventilation: median tidal volume measured with the pneumotachometer (9.5 mL/kg [interquartile range, 8.2-11.7 mL/kg]) was significantly higher than tidal volume measured either at the ventilator (8.2 mL/kg [7.1-9.6 mL/kg]) or at the proximal flow sensor (8.1 mL/kg [7.2-10.0 mL/kg]) (p tidal volume measured with the pneumotachometer (10.2 mL/kg [8.8-12.4 mL/kg]) was significantly higher than tidal volume measured either at the ventilator (8.0 mL/kg [7.1-9.7 mL/kg]) or at the proximal flow sensor (8.5 mL/kg [7.3-10.4 mL/kg]) (p Tidal volume measured either at the endotracheal tube with a proximal flow sensor or at the ventilator with compensation for tubing compliance are both significantly lower than tidal volume measured with a calibrated pneumotachometer. This underestimation of delivered tidal volume may be particularly important when managing children with acute respiratory distress syndrome.

  13. CYLINDER-BASED SELF-CALIBRATION OF A PANORAMIC TERRESTRIAL LASER SCANNER

    Directory of Open Access Journals (Sweden)

    T. O. Chan

    2012-07-01

    Full Text Available Terrestrial Laser Scanners (TLSs have become state-of-the-art metrological sensors for many surveying purposes in recent years. Due to the demand for high precision surveying with TLSs, efficient, rigorous and in-situ calibration methodologies are always desired. Recent research on in-situ calibration with planar features has demonstrated improved cost-effectiveness and promising results (Glennie and Lichti, 2010; Chow et al., 2011; Chow et al., 2012. However, if there is a need for calibrating the scanners when sufficient plane surfaces with several orientations are not available, as commonly occurs indoors, other common geometric features, namely cylindrical structures, can be used as alternative geometric constraints for in-situ self-calibration. Cylindrical features can be found in indoor environments such as water pipes attached to the walls or suspended from ceilings, concrete pillars, metal poles and many others. In this paper, three 3D models of cylinders, with vertical and horizontal orientations containing one scaling, two rotational and two translational parameters are discussed. The cylinder models are parameterized with the sexternal orientation parameters and the additional parameters as the least-squares functional models for the self-calibration. The selfcalibration is examined with the real data obtained from the Lecia HDS6100 panoramic TLS. The results of vertical, horizontal and mixed cylinder-based calibration with data captured by different scanner position are analysed in detail in terms of the parameters correlations. The results show realistic estimation of calibration parameters for several cases. The results also suggest that using both vertical and horizontal cylinders for the calibration can effectively decorrelate the parameters especially for the case of lack of cylinder point cloud overlap. The concepts developed in this paper might also be extended to the hybrid type TLSs, as well as to the self-calibration of

  14. European Prospective Investigation into Cancer and Nutrition (EPIC) calibration study: rationale, design and population characteristics

    DEFF Research Database (Denmark)

    Slimani, N.; Kaaks, R.; Ferrari, P.

    2002-01-01

    The European Prospective Investigation into Cancer and Nutrition (EPIC), which covers a large cohort of half a million men and women from 23 European centres in 10 Western European countries, was designed to study the relationship between diet and the risk of chronic diseases, particularly cancer......, a calibration approach was developed. This approach involved an additional dietary assessment common across study populations to re-express individual dietary intakes according to the same reference scale. A single 24-hour diet recall was therefore collected, as the EPIC reference calibration method, from...... in a large multi-centre European study. These studies showed that, despite certain inherent methodological and logistic constraints, a study design such as this one works relatively well in practice. The average response in the calibration study was 78.3% and ranged from 46.5% to 92.5%. The calibration...

  15. Meteorological Sensor Array (MSA)-Phase I. Volume 3 (Pre-Field Campaign Sensor Calibration)

    Science.gov (United States)

    2015-07-01

    Phase I thermodynamic sensors (barometers, thermometers, hygrometers, and pyranometers ). This report documents the results of a detailed calibration...barometer, thermometer, hydrometer, pyranometer 16. SECURITY CLASSIFICATION OF: 17. LIMITATION OF ABSTRACT UU 18. NUMBER...60 (02 Mar); the logger associated with each non-standard RH sensor is demonstrated in the legend ...............21 Fig. 16 Pyranometers percent

  16. Cuban-Brazilian comparison of the calibration procedures for surface contamination monitors

    International Nuclear Information System (INIS)

    Salas, Gonzalo W.; Garcia, Jose A.T.; Ramos, Manoel M.O.

    2007-01-01

    The calibration laboratory of the CPHR, Centro de Proteccion e Higiene de las Radiaciones, is the only laboratory that performs calibrations of radiation protection instruments in Cuba. Recently they started a cooperation project with IAEA, code number CUB3002-01 with the task title 'Assuring and Demonstrating the safety of radioactive Waste Management'. Within the frame of this project they have a compromise to implement the calibration service for surface contamination monitors, for which they received instruments and sources. In 2006 the Laboratorio Nacional de Metrologia das Radiacoes Ionizantes (LNMRI) from the Instituto de Radioprotecao e Dosimetria, IRD/CNEN, was invited to perform a mission to CPHR in order to evaluate the level of implementation of the service trough the insertion of the service in the Quality Assurance System of their laboratory, and perform a comparison exercise to test the calibration method implemented. The results were evaluated by the methodology recommended in the ISO/IEC 43-1 standard. These results show that all values from both laboratories are comparable for most radionuclides except for Cl-36. The preliminary investigation about this situation is presented in the paper. (author)

  17. Using MCNP for in-core instrument calibration in CANDU

    Energy Technology Data Exchange (ETDEWEB)

    Taylor, D.C. [Point Lepreau Generating Station, NB Power, Lepreau, New Brunswick (Canada); Anghel, V.N.P.; Sur, B. [Atomic Energy of Canada Limited, Chalk River, Ontario (Canada)

    2002-07-01

    The calibration of in-core instruments is important for safe and economical CANDU operation. However, in-core detectors are not normally suited to bench calibration procedures. This paper describes the use and validation of detailed neutron transport calculations for the purpose of calibrating the response of in-core neutron flux detectors. The Monte-Carlo transport code, MCNP, was used to model the thermal neutron flux distribution in the region around self-powered in-core flux detectors (ICFDs), and in the vicinity of the calandria edge. The ICFD model was used to evaluate the reduction in signal of a given detector (the 'detector shading factor') due to neutron absorption in surrounding materials, detectors, and lead-cables. The calandria edge model was used to infer the accuracy of the calandria edge position from flux scans performed by AECL's traveling flux detector (TFD) system. The MCNP results were checked against experimental results on ICFDs, and also against shading factors computed by other means. The use of improved in-core detector calibration factors obtained by this new methodology will improve the accuracy of spatial flux control performance in CANDU-6 reactors. The accurate determination of TFD based calandria edge position is useful in the quantitative measurement of changes in in-core component dimensions and position due to aging, such as pressure tube sag. (author)

  18. Implied Volatility Surface: Construction Methodologies and Characteristics

    OpenAIRE

    Cristian Homescu

    2011-01-01

    The implied volatility surface (IVS) is a fundamental building block in computational finance. We provide a survey of methodologies for constructing such surfaces. We also discuss various topics which can influence the successful construction of IVS in practice: arbitrage-free conditions in both strike and time, how to perform extrapolation outside the core region, choice of calibrating functional and selection of numerical optimization algorithms, volatility surface dynamics and asymptotics.

  19. Exposure-rate calibration using large-area calibration pads

    International Nuclear Information System (INIS)

    Novak, E.F.

    1988-09-01

    The US Department of Energy (DOE) Office of Remedial Action and Waste Technology established the Technical Measurements Center (TMC) at the DOE Grand Junction Projects Office (GJPO) in Grand Junction, Colorado, to standardize, calibrate, and compare measurements made in support of DOE remedial action programs. A set of large-area, radioelement-enriched concrete pads was constructed by the DOE in 1978 at the Walker Field Airport in Grand Junction for use as calibration standards for airborne gamma-ray spectrometer systems. The use of these pads was investigated by the TMC as potential calibration standards for portable scintillometers employed in measuring gamma-ray exposure rates at Uranium Mill Tailings Remedial Action (UMTRA) project sites. Data acquired on the pads using a pressurized ionization chamber (PIC) and three scintillometers are presented as an illustration of an instrumental calibration. Conclusions and recommended calibration procedures are discussed, based on the results of these data

  20. Comparison of CT number calibration techniques for CBCT-based dose calculation

    International Nuclear Information System (INIS)

    Dunlop, Alex; McQuaid, Dualta; Nill, Simeon; Hansen, Vibeke N.; Oelfke, Uwe; Murray, Julia; Bhide, Shreerang; Harrington, Kevin; Poludniowski, Gavin; Nutting, Christopher; Newbold, Kate

    2015-01-01

    The aim of this work was to compare and validate various computed tomography (CT) number calibration techniques with respect to cone beam CT (CBCT) dose calculation accuracy. CBCT dose calculation accuracy was assessed for pelvic, lung, and head and neck (H and N) treatment sites for two approaches: (1) physics-based scatter correction methods (CBCT r ); (2) density override approaches including assigning water density to the entire CBCT (W), assignment of either water or bone density (WB), and assignment of either water or lung density (WL). Methods for CBCT density assignment within a commercially available treatment planning system (RS auto ), where CBCT voxels are binned into six density levels, were assessed and validated. Dose-difference maps and dose-volume statistics were used to compare the CBCT dose distributions with the ground truth of a planning CT acquired the same day as the CBCT. For pelvic cases, all CTN calibration methods resulted in average dose-volume deviations below 1.5 %. RS auto provided larger than average errors for pelvic treatments for patients with large amounts of adipose tissue. For H and N cases, all CTN calibration methods resulted in average dose-volume differences below 1.0 % with CBCT r (0.5 %) and RS auto (0.6 %) performing best. For lung cases, WL and RS auto methods generated dose distributions most similar to the ground truth. The RS auto density override approach is an attractive option for CTN adjustments for a variety of anatomical sites. RS auto methods were validated, resulting in dose calculations that were consistent with those calculated on diagnostic-quality CT images, for CBCT images acquired of the lung, for patients receiving pelvic RT in cases without excess adipose tissue, and for H and N cases. (orig.) [de

  1. Improved USGS methodology for assessing continuous petroleum resources

    Science.gov (United States)

    Charpentier, Ronald R.; Cook, Troy A.

    2010-01-01

    This report presents an improved methodology for estimating volumes of continuous (unconventional) oil and gas resources within the United States and around the world. The methodology is based on previously developed U.S. Geological Survey methodologies that rely on well-scale production data. Improvements were made primarily to how the uncertainty about estimated ultimate recoveries is incorporated in the estimates. This is particularly important when assessing areas with sparse or no production data, because the new methodology allows better use of analog data from areas with significant discovery histories.

  2. Improvements in absolute seismometer sensitivity calibration using local earth gravity measurements

    Science.gov (United States)

    Anthony, Robert E.; Ringler, Adam; Wilson, David

    2018-01-01

    The ability to determine both absolute and relative seismic amplitudes is fundamentally limited by the accuracy and precision with which scientists are able to calibrate seismometer sensitivities and characterize their response. Currently, across the Global Seismic Network (GSN), errors in midband sensitivity exceed 3% at the 95% confidence interval and are the least‐constrained response parameter in seismic recording systems. We explore a new methodology utilizing precise absolute Earth gravity measurements to determine the midband sensitivity of seismic instruments. We first determine the absolute sensitivity of Kinemetrics EpiSensor accelerometers to 0.06% at the 99% confidence interval by inverting them in a known gravity field at the Albuquerque Seismological Laboratory (ASL). After the accelerometer is calibrated, we install it in its normal configuration next to broadband seismometers and subject the sensors to identical ground motions to perform relative calibrations of the broadband sensors. Using this technique, we are able to determine the absolute midband sensitivity of the vertical components of Nanometrics Trillium Compact seismometers to within 0.11% and Streckeisen STS‐2 seismometers to within 0.14% at the 99% confidence interval. The technique enables absolute calibrations from first principles that are traceable to National Institute of Standards and Technology (NIST) measurements while providing nearly an order of magnitude more precision than step‐table calibrations.

  3. Model validation and calibration based on component functions of model output

    International Nuclear Information System (INIS)

    Wu, Danqing; Lu, Zhenzhou; Wang, Yanping; Cheng, Lei

    2015-01-01

    The target in this work is to validate the component functions of model output between physical observation and computational model with the area metric. Based on the theory of high dimensional model representations (HDMR) of independent input variables, conditional expectations are component functions of model output, and the conditional expectations reflect partial information of model output. Therefore, the model validation of conditional expectations tells the discrepancy between the partial information of the computational model output and that of the observations. Then a calibration of the conditional expectations is carried out to reduce the value of model validation metric. After that, a recalculation of the model validation metric of model output is taken with the calibrated model parameters, and the result shows that a reduction of the discrepancy in the conditional expectations can help decrease the difference in model output. At last, several examples are employed to demonstrate the rationality and necessity of the methodology in case of both single validation site and multiple validation sites. - Highlights: • A validation metric of conditional expectations of model output is proposed. • HDRM explains the relationship of conditional expectations and model output. • An improved approach of parameter calibration updates the computational models. • Validation and calibration process are applied at single site and multiple sites. • Validation and calibration process show a superiority than existing methods

  4. Determination of fragrance content in perfume by Raman spectroscopy and multivariate calibration

    Science.gov (United States)

    Godinho, Robson B.; Santos, Mauricio C.; Poppi, Ronei J.

    2016-03-01

    An alternative methodology is herein proposed for determination of fragrance content in perfumes and their classification according to the guidelines established by fine perfume manufacturers. The methodology is based on Raman spectroscopy associated with multivariate calibration, allowing the determination of fragrance content in a fast, nondestructive, and sustainable manner. The results were considered consistent with the conventional method, whose standard error of prediction values was lower than the 1.0%. This result indicates that the proposed technology is a feasible analytical tool for determination of the fragrance content in a hydro-alcoholic solution for use in manufacturing, quality control and regulatory agencies.

  5. Model calibration for building energy efficiency simulation

    International Nuclear Information System (INIS)

    Mustafaraj, Giorgio; Marini, Dashamir; Costa, Andrea; Keane, Marcus

    2014-01-01

    Highlights: • Developing a 3D model relating to building architecture, occupancy and HVAC operation. • Two calibration stages developed, final model providing accurate results. • Using an onsite weather station for generating the weather data file in EnergyPlus. • Predicting thermal behaviour of underfloor heating, heat pump and natural ventilation. • Monthly energy saving opportunities related to heat pump of 20–27% was identified. - Abstract: This research work deals with an Environmental Research Institute (ERI) building where an underfloor heating system and natural ventilation are the main systems used to maintain comfort condition throughout 80% of the building areas. Firstly, this work involved developing a 3D model relating to building architecture, occupancy and HVAC operation. Secondly, the calibration methodology, which consists of two levels, was then applied in order to insure accuracy and reduce the likelihood of errors. To further improve the accuracy of calibration a historical weather data file related to year 2011, was created from the on-site local weather station of ERI building. After applying the second level of calibration process, the values of Mean bias Error (MBE) and Cumulative Variation of Root Mean Squared Error (CV(RMSE)) on hourly based analysis for heat pump electricity consumption varied within the following ranges: (MBE) hourly from −5.6% to 7.5% and CV(RMSE) hourly from 7.3% to 25.1%. Finally, the building was simulated with EnergyPlus to identify further possibilities of energy savings supplied by a water to water heat pump to underfloor heating system. It found that electricity consumption savings from the heat pump can vary between 20% and 27% on monthly bases

  6. Comparison and uncertainty evaluation of different calibration protocols and ionization chambers for low-energy surface brachytherapy dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Candela-Juan, C., E-mail: ccanjuan@gmail.com [Radiation Oncology Department, La Fe University and Polytechnic Hospital, Valencia 46026 (Spain); Vijande, J. [Department of Atomic, Molecular, and Nuclear Physics, University of Valencia, Burjassot 46100, Spain and Instituto de Física Corpuscular (UV-CSIC), Paterna 46980 (Spain); García-Martínez, T. [Radiation Oncology Department, Hospital La Ribera, Alzira 46600 (Spain); Niatsetski, Y.; Nauta, G.; Schuurman, J. [Elekta Brachytherapy, Veenendaal 3905 TH (Netherlands); Ouhib, Z. [Radiation Oncology Department, Lynn Regional Cancer Center, Boca Raton Community Hospital, Boca Raton, Florida 33486 (United States); Ballester, F. [Department of Atomic, Molecular, and Nuclear Physics, University of Valencia, Burjassot 46100 (Spain); Perez-Calatayud, J. [Radiation Oncology Department, La Fe University and Polytechnic Hospital, Valencia 46026, Spain and Department of Radiotherapy, Clínica Benidorm, Benidorm 03501 (Spain)

    2015-08-15

    Purpose: A surface electronic brachytherapy (EBT) device is in fact an x-ray source collimated with specific applicators. Low-energy (<100 kVp) x-ray beam dosimetry faces several challenges that need to be addressed. A number of calibration protocols have been published for x-ray beam dosimetry. The media in which measurements are performed are the fundamental difference between them. The aim of this study was to evaluate the surface dose rate of a low-energy x-ray source with small field applicators using different calibration standards and different small-volume ionization chambers, comparing the values and uncertainties of each methodology. Methods: The surface dose rate of the EBT unit Esteya (Elekta Brachytherapy, The Netherlands), a 69.5 kVp x-ray source with applicators of 10, 15, 20, 25, and 30 mm diameter, was evaluated using the AAPM TG-61 (based on air kerma) and International Atomic Energy Agency (IAEA) TRS-398 (based on absorbed dose to water) dosimetry protocols for low-energy photon beams. A plane parallel T34013 ionization chamber (PTW Freiburg, Germany) calibrated in terms of both absorbed dose to water and air kerma was used to compare the two dosimetry protocols. Another PTW chamber of the same model was used to evaluate the reproducibility between these chambers. Measurements were also performed with two different Exradin A20 (Standard Imaging, Inc., Middleton, WI) chambers calibrated in terms of air kerma. Results: Differences between surface dose rates measured in air and in water using the T34013 chamber range from 1.6% to 3.3%. No field size dependence has been observed. Differences are below 3.7% when measurements with the A20 and the T34013 chambers calibrated in air are compared. Estimated uncertainty (with coverage factor k = 1) for the T34013 chamber calibrated in water is 2.2%–2.4%, whereas it increases to 2.5% and 2.7% for the A20 and T34013 chambers calibrated in air, respectively. The output factors, measured with the PTW chambers

  7. Correction factors of commercial radionuclide calibrators for several measurement geometries of radiopharmaceuticals

    International Nuclear Information System (INIS)

    Correia, Amanda Ribeiro

    2011-01-01

    In order to reach therapy and diagnosis objectives, the activity must be determined with high accuracy to administer a radiopharmaceutical to a patient. Initially, a glass vial with the radiopharmaceutical is placed into the radionuclide calibrator to determine its activity. Subsequently, an aliquot is transferred to a syringe and again its activity is measured on the calibrator before being administered to the patient. The glass vial and the syringe are different in many aspects as the calibration factors too, which may cause incorrect activities administered to the patient. This study aims to determine the correction factors, as well as the values of the uncertainties associated to two distinct models of calibrators: one that uses ionization chamber and another Geiger-Mueller as detectors. The radionuclides chosen were 99 Tc m and 123 1 and the containers were glass vials (type lOR and P6) and plastic syringes of 3 and 5 mL. The correction factors for each type of vials or syringe were determined as a function of volume and type of calibrator. Activity measurements comparison was also made involving several radionuclide calibrators of different models belonging to four nuclear medicine hospitals and to National Metrology Laboratory of lionizing Radiation (LNMRI). In the measurements of activity values larger than allowed by CNEN NN-3.05 norm, results have shown deviations for syringes in calibrator with Geiger-Mueller detectors and for both radionuclides. (author)

  8. Establishment of Accurate Calibration Curve for National Verification at a Large Scale Input Accountability Tank in RRP - For Strengthening State System for Meeting Safeguards Obligation

    International Nuclear Information System (INIS)

    Goto, Y.; Kato, T.; Nidaira, K.

    2010-01-01

    Tanks are installed in a reprocessing plant for spent fuel in order to account solution of nuclear material. The careful measurement of volume in tanks is crucial to implement accurate accounting of nuclear material. The calibration curve related with the volume and level of solution needs to be constructed, where the level is determined by differential pressure of dip tubes in tanks. More than one calibration curves depending on the height are commonly applied for each tank, but it's not explicitly decided how many segments are used, where to select segment, or what order of polynomial curve. Here we present the rational construction technique of giving optimum calibration curves and their characteristics. The tank calibration work has been conducted in the course of contract with Japan Safeguards Office (JSGO) about safeguards information treatment. (author)

  9. Calibration of the BASS acoustic current meter with carrageenan agar

    Science.gov (United States)

    Morrison, A.T.; Williams, A.J.; Martini, M.

    1993-01-01

    The BASS current meter can measure currents down to the millimeter per second range. Due to the dependence of zero offset on pressure, determining a sensor referenced velocity requires accurate in situ zeroing of the meter. Previously, flow was restricted during calibration by placing plastic bags around the acoustic volume. In this paper, bacterial grade and carrageenan agars are used in the laboratory to create a zero flow condition during calibration and are shown to be acoustically transparent. Additionally, the results of open ocean and dockside carrageenan and plastic bag comparisons are presented. Carrageenan is shown to reliably provide a low noise, zero mean flow environment that is largely independent of ambient conditions. The improved zeros make millimeter per second accuracy possible under field conditions.

  10. A calibration rig for multi-component internal strain gauge balance using the new design-of-experiment (DOE) approach

    Science.gov (United States)

    Nouri, N. M.; Mostafapour, K.; Kamran, M.

    2018-02-01

    In a closed water-tunnel circuit, the multi-component strain gauge force and moment sensor (also known as balance) are generally used to measure hydrodynamic forces and moments acting on scaled models. These balances are periodically calibrated by static loading. Their performance and accuracy depend significantly on the rig and the method of calibration. In this research, a new calibration rig was designed and constructed to calibrate multi-component internal strain gauge balances. The calibration rig has six degrees of freedom and six different component-loading structures that can be applied separately and synchronously. The system was designed based on the applicability of formal experimental design techniques, using gravity for balance loading and balance positioning and alignment relative to gravity. To evaluate the calibration rig, a six-component internal balance developed by Iran University of Science and Technology was calibrated using response surface methodology. According to the results, calibration rig met all design criteria. This rig provides the means by which various methods of formal experimental design techniques can be implemented. The simplicity of the rig saves time and money in the design of experiments and in balance calibration while simultaneously increasing the accuracy of these activities.

  11. Pilot study to verify the calibration of electrometers

    International Nuclear Information System (INIS)

    Becker, P.; Meghzifene, A.

    2002-01-01

    National Laboratory for Electrical Measurements has not yet developed its capability for the standardization of small electrical charge produced by DC, the IRD is trying to verify its standardization procedures of the electrical charge through a comparison programme. This subject was discussed with a major electrometer manufacturer that has offered to provide free of charge, three of their electrometer calibration standards for a pilot run. The model to be provided consists of four calibrated resistors and two calibrated capacitors, covering the charge/current range of interest. For producing charge or current a standard DC voltage must be applied to these components. Since practically all-modern electrometers measure using virtual ground, this methodology is viable. The IRD, in collaboration with the IAEA, wishes to invite interested laboratories to participate in this pilot comparison programme. This exercise is expected to be useful for all participants and will hopefully open the way for the establishment of routine comparisons in this area. The results will be discussed and published in an appropriate journal. Interested institutions should contact directly Mr. Paulo H. B. Becker through e-mail (pbecker at ird.gov.br) or fax +55 21 24421950 informing him of the model and manufacturer of the electrometer to be used for the pilot study and discuss all practical details. (author)

  12. Dosimetric methodology and reference system for diagnostic level X radiation

    International Nuclear Information System (INIS)

    Potiens, Maria da Penha Albuquerque

    1999-01-01

    Several methodologies for the calibration of diagnostic radiology instruments were developed and established at the Calibration Laboratory of IPEN. These established radiation qualities are recommended by international standards. The methods may be used in the calibration procedures of survey meters used in radiation protection measurements (scattered radiation), instruments used in direct beams (attenuated and non attenuated beams) and quality control instruments. A reference system was proposed using two identical ionization chambers developed at IPEN. They differ only by the collecting electrode material, one of aluminium and the other of graphite. The different energetic dependence of the chamber's response provided a ratio related to the tube potential. The variation of only 0.28%, from 14.3 to 111 keV, on the energetic dependence of the graphite electrode chamber, provided the possibility of air kerma rate determination in the studied radiation beams. (author)

  13. Plasma volume methodology: Evans blue, hemoglobin-hematocrit, and mass density transformations

    Science.gov (United States)

    Greenleaf, J. E.; Hinghofer-Szalkay, H.

    1985-01-01

    Methods for measuring absolute levels and changes in plasma volume are presented along with derivations of pertinent equations. Reduction in variability of the Evans blue dye dilution technique using chromatographic column purification suggests that the day-to-day variability in the plasma volume in humans is less than + or - 20 m1. Mass density determination using the mechanical-oscillator technique provides a method for measuring vascular fluid shifts continuously for assessing the density of the filtrate, and for quantifying movements of protein across microvascular walls. Equations for the calculation of volume and density of shifted fluid are presented.

  14. Calibration and Validation Parameter of Hydrologic Model HEC-HMS using Particle Swarm Optimization Algorithms – Single Objective

    Directory of Open Access Journals (Sweden)

    R. Garmeh

    2016-02-01

    model that simulates both wet and dry weatherbehavior.Programming of HEC –HMS has been done by MATLAB and techniques such as elite mutation and creating confusion have been used in order to strengthen the algorithm and improve the results. The event-based HEC-HMS model simulatesthe precipitation-runoff process for each set of parameter values generated by PSO. Turbulentand elitism with mutation are also employed to deal with PSO premature convergence. The integrated PSO-HMS model is tested on the Kardeh dam basin located in the Khorasan Razavi province. Results and Discussion: Input parameters of hydrologic models are seldomknown with certainty. Therefore, they are not capable ofdescribing the exact hydrologic processes. Input data andstructural uncertainties related to scale and approximationsin system processes are different sources of uncertainty thatmake it difficult to model exact hydrologic phenomena.In automatic calibration, the parameter values dependon the objective function of the search or optimization algorithm.In characterizing a runoff hydrograph, threecharacteristics of time-to-peak, peak of discharge and totalrunoff volume are of the most importance. It is thereforeimportant that we simulate and observe hydrographs matchas much as possible in terms of those characteristics. Calibration was carried out in single objective cases. Model calibration in single-objective approach with regard to the objective function in the event of NASH and RMSE were conducted separately.The results indicated that the capability of the model was calibrated to an acceptable level of events. Continuing calibration results were evaluated by four different criteria.Finally, to validate the model parameters with those obtained from the calibration, tests perfomed indicated poor results. Although, based on the calibration and verification of individual events one event remains, suggesting set is a possible parameter. Conclusion: All events were evaluated by validations and the

  15. Density and volume measurements of reprocessing plant feed

    International Nuclear Information System (INIS)

    Platzer, R.; Carrier, M.; Neuilly, M.; Dedaldechamp, P.

    1985-05-01

    A theoretical study of the phenomenon of gas bubbles formation within a liquid led to an adaptation of the differential pressure bubbling technique for the measurement of liquid levels and densities in tanks. Experiments, carried out on a 800 liters tank with water and uranyl nitrate solutions had the double aim to study the precision attainable on volume and density measurements and to design a method for corrections of influencing factors. In parallel, procedures for transfer of known volumes through the use of siphons and for tank calibration by liquid level measurement are also investigated. The paper presents the first results obtained so far and the conclusions to be drawn for the elaboration of calibration and exploitation procedures suitables for use in reprocessing plants. The demonstration to transfer mass of solution with an accuracy of 0.1% is made [fr

  16. Automatic camera to laser calibration for high accuracy mobile mapping systems using INS

    Science.gov (United States)

    Goeman, Werner; Douterloigne, Koen; Gautama, Sidharta

    2013-09-01

    A mobile mapping system (MMS) is a mobile multi-sensor platform developed by the geoinformation community to support the acquisition of huge amounts of geodata in the form of georeferenced high resolution images and dense laser clouds. Since data fusion and data integration techniques are increasingly able to combine the complementary strengths of different sensor types, the external calibration of a camera to a laser scanner is a common pre-requisite on today's mobile platforms. The methods of calibration, nevertheless, are often relatively poorly documented, are almost always time-consuming, demand expert knowledge and often require a carefully constructed calibration environment. A new methodology is studied and explored to provide a high quality external calibration for a pinhole camera to a laser scanner which is automatic, easy to perform, robust and foolproof. The method presented here, uses a portable, standard ranging pole which needs to be positioned on a known ground control point. For calibration, a well studied absolute orientation problem needs to be solved. In many cases, the camera and laser sensor are calibrated in relation to the INS system. Therefore, the transformation from camera to laser contains the cumulated error of each sensor in relation to the INS. Here, the calibration of the camera is performed in relation to the laser frame using the time synchronization between the sensors for data association. In this study, the use of the inertial relative movement will be explored to collect more useful calibration data. This results in a better intersensor calibration allowing better coloring of the clouds and a more accurate depth mask for images, especially on the edges of objects in the scene.

  17. Recycle operations as a methodology for radioactive waste volume reduction

    International Nuclear Information System (INIS)

    Rasmussen, G.A.

    1985-01-01

    The costs for packaging, transportation and burial of low-level radioactive metallic waste have become so expensive that an alternate method of decontamination for volume reduction prior to disposal can now be justified. The operation of a large-scale centralized recycle center for decontamination of selected low level radioactive waste has been proven to be an effective method for waste volume reduction and for retrieving valuable materials for unlimited use. The centralized recycle center concept allows application of state-of-the-art decontamination technology resulting in a reduction in utility disposal costs and a reduction in overall net amount of material being buried. Examples of specific decontamination process activities at the centralized facility will be reviewed along with a discussion of the economic impact of decontamination for recycling and volume reduction. Based on almost two years of operation of a centralized decontamination facility, a demonstrated capability exists. The concept has been cost effective and proves that valuable resources can be recycled

  18. Reactor analysis support package (RASP). Volume 7. PWR set-point methodology. Final report

    International Nuclear Information System (INIS)

    Temple, S.M.; Robbins, T.R.

    1986-09-01

    This report provides an overview of the basis and methodology requirements for determining Pressurized Water Reactor (PWR) technical specifications related setpoints and focuses on development of the methodology for a reload core. Additionally, the report documents the implementation and typical methods of analysis used by PWR vendors during the 1970's to develop Protection System Trip Limits (or Limiting Safety System Settings) and Limiting Conditions for Operation. The descriptions of the typical setpoint methodologies are provided for Nuclear Steam Supply Systems as designed and supplied by Babcock and Wilcox, Combustion Engineering, and Westinghouse. The description of the methods of analysis includes the discussion of the computer codes used in the setpoint methodology. Next, the report addresses the treatment of calculational and measurement uncertainties based on the extent to which such information was available for each of the three types of PWR. Finally, the major features of the setpoint methodologies are compared, and the principal effects of each particular methodology on plant operation are summarized for each of the three types of PWR

  19. Postal auditing methodology used to find out the performance of high rate brachytherapy equipment

    International Nuclear Information System (INIS)

    Morales, J.A.; Campa, R.

    1998-01-01

    This work describes results from a methodology implemented at the Secondary Laboratory for Dosimetric Calibration at CPHR used to check the brachytherapy performance at high doses rate using Cesium 137 or cobalt 60 sources

  20. Mapping soil deformation around plant roots using in vivo 4D X-ray Computed Tomography and Digital Volume Correlation.

    Science.gov (United States)

    Keyes, S D; Gillard, F; Soper, N; Mavrogordato, M N; Sinclair, I; Roose, T

    2016-06-14

    The mechanical impedance of soils inhibits the growth of plant roots, often being the most significant physical limitation to root system development. Non-invasive imaging techniques have recently been used to investigate the development of root system architecture over time, but the relationship with soil deformation is usually neglected. Correlative mapping approaches parameterised using 2D and 3D image data have recently gained prominence for quantifying physical deformation in composite materials including fibre-reinforced polymers and trabecular bone. Digital Image Correlation (DIC) and Digital Volume Correlation (DVC) are computational techniques which use the inherent material texture of surfaces and volumes, captured using imaging techniques, to map full-field deformation components in samples during physical loading. Here we develop an experimental assay and methodology for four-dimensional, in vivo X-ray Computed Tomography (XCT) and apply a Digital Volume Correlation (DVC) approach to the data to quantify deformation. The method is validated for a field-derived soil under conditions of uniaxial compression, and a calibration study is used to quantify thresholds of displacement and strain measurement. The validated and calibrated approach is then demonstrated for an in vivo test case in which an extending maize root in field-derived soil was imaged hourly using XCT over a growth period of 19h. This allowed full-field soil deformation data and 3D root tip dynamics to be quantified in parallel for the first time. This fusion of methods paves the way for comparative studies of contrasting soils and plant genotypes, improving our understanding of the fundamental mechanical processes which influence root system development. Copyright © 2016 Elsevier Ltd. All rights reserved.

  1. A glimpse beneath Antarctic sea ice: observation of platelet-layer thickness and ice-volume fraction with multi-frequency EM

    Science.gov (United States)

    Hendricks, S.; Hoppmann, M.; Hunkeler, P. A.; Kalscheuer, T.; Gerdes, R.

    2015-12-01

    In Antarctica, ice crystals (platelets) form and grow in supercooled waters below ice shelves. These platelets rise and accumulate beneath nearby sea ice to form a several meter thick sub-ice platelet layer. This special ice type is a unique habitat, influences sea-ice mass and energy balance, and its volume can be interpreted as an indicator for ice - ocean interactions. Although progress has been made in determining and understanding its spatio-temporal variability based on point measurements, an investigation of this phenomenon on a larger scale remains a challenge due to logistical constraints and a lack of suitable methodology. In the present study, we applied a lateral constrained Marquardt-Levenberg inversion to a unique multi-frequency electromagnetic (EM) induction sounding dataset obtained on the ice-shelf influenced fast-ice regime of Atka Bay, eastern Weddell Sea. We adapted the inversion algorithm to incorporate a sensor specific signal bias, and confirmed the reliability of the algorithm by performing a sensitivity study using synthetic data. We inverted the field data for sea-ice and sub-ice platelet-layer thickness and electrical conductivity, and calculated ice-volume fractions from platelet-layer conductivities using Archie's Law. The thickness results agreed well with drill-hole validation datasets within the uncertainty range, and the ice-volume fraction also yielded plausible results. Our findings imply that multi-frequency EM induction sounding is a suitable approach to efficiently map sea-ice and platelet-layer properties. However, we emphasize that the successful application of this technique requires a break with traditional EM sensor calibration strategies due to the need of absolute calibration with respect to a physical forward model.

  2. Study of the performance of diagnostic radiology instruments during calibration

    International Nuclear Information System (INIS)

    Freitas, Rodrigo N. de; Vivolo, Vitor; Potiens, Maria da Penha A.

    2008-01-01

    Full text: The instruments used in diagnostic radiology measurements represent 8 % of the tested instruments by the calibration laboratory of IPEN annually (approximately 1600 in 2007). Considering that the calibration of this kind of instrument is performed biannually it is possible to conclude that almost 300 instruments are being used to measure the air kerma in diagnostic radiology clinics to determine the in beam values (in front of the patient), attenuated measurements (behind the patient) and scattered radiation. This work presents the results of the calibration of the instruments used in mammography, computed tomography, dental and conventional diagnostic radiology dosimetry, performed during the period of 2005 to 2007. Their performances during the calibrations measurements were evaluated. Although at the calibration laboratory there are three available series of radiation quality to this type of calibration (RQR, N and M, according to standards IEC 61267 and ISO 4037-1.), the applications can be assorted (general radiology, computed tomography, mammography, radiation protection and fluoroscopy). Depending on its design and behaviour , one kind of instrument can be used for one or more type of applications. The instruments normally used for diagnostic radiology measurements are ionization chambers with volumes varying from 3 to 1800 cm 3 , and can be cylindrical, spherical or plane parallel plates kind. They usually are sensitive to photon particles, with energies greater than 15 keV and can be used up to 1200 keV. In this work they were tested in X radiation fields from 25 to 150 kV, in specific qualities depending on the utilization of the instrument. The calibration results of 390 instruments received from 2005 to 2007 were analyzed. About 20 instruments were not able to be calibrated due to bad functioning. The calibration coefficients obtained were between 0.88 and 1.24. The uncertainties were always less than ± 3.6% to instruments used in scattered

  3. CREPT-MCNP code for efficiency calibration of HPGe detectors with the representative point method.

    Science.gov (United States)

    Saegusa, Jun

    2008-01-01

    The representative point method for the efficiency calibration of volume samples has been previously proposed. For smoothly implementing the method, a calculation code named CREPT-MCNP has been developed. The code estimates the position of a representative point which is intrinsic to each shape of volume sample. The self-absorption correction factors are also given to make correction on the efficiencies measured at the representative point with a standard point source. Features of the CREPT-MCNP code are presented.

  4. Reference Materials for Calibration of Analytical Biases in Quantification of DNA Methylation.

    Science.gov (United States)

    Yu, Hannah; Hahn, Yoonsoo; Yang, Inchul

    2015-01-01

    Most contemporary methods for the quantification of DNA methylation employ bisulfite conversion and PCR amplification. However, many reports have indicated that bisulfite-mediated PCR methodologies can result in inaccurate measurements of DNA methylation owing to amplification biases. To calibrate analytical biases in quantification of gene methylation, especially those that arise during PCR, we utilized reference materials that represent exact bisulfite-converted sequences with 0% and 100% methylation status of specific genes. After determining relative quantities using qPCR, pairs of plasmids were gravimetrically mixed to generate working standards with predefined DNA methylation levels at 10% intervals in terms of mole fractions. The working standards were used as controls to optimize the experimental conditions and also as calibration standards in melting-based and sequencing-based analyses of DNA methylation. Use of the reference materials enabled precise characterization and proper calibration of various biases during PCR and subsequent methylation measurement processes, resulting in accurate measurements.

  5. Advanced piloted aircraft flight control system design methodology. Volume 1: Knowledge base

    Science.gov (United States)

    Mcruer, Duane T.; Myers, Thomas T.

    1988-01-01

    The development of a comprehensive and electric methodology for conceptual and preliminary design of flight control systems is presented and illustrated. The methodology is focused on the design stages starting with the layout of system requirements and ending when some viable competing system architectures (feedback control structures) are defined. The approach is centered on the human pilot and the aircraft as both the sources of, and the keys to the solution of, many flight control problems. The methodology relies heavily on computational procedures which are highly interactive with the design engineer. To maximize effectiveness, these techniques, as selected and modified to be used together in the methodology, form a cadre of computational tools specifically tailored for integrated flight control system preliminary design purposes. While theory and associated computational means are an important aspect of the design methodology, the lore, knowledge and experience elements, which guide and govern applications are critical features. This material is presented as summary tables, outlines, recipes, empirical data, lists, etc., which encapsulate a great deal of expert knowledge. Much of this is presented in topical knowledge summaries which are attached as Supplements. The composite of the supplements and the main body elements constitutes a first cut at a a Mark 1 Knowledge Base for manned-aircraft flight control.

  6. The cost of uniqueness in groundwater model calibration

    Science.gov (United States)

    Moore, Catherine; Doherty, John

    2006-04-01

    Calibration of a groundwater model requires that hydraulic properties be estimated throughout a model domain. This generally constitutes an underdetermined inverse problem, for which a solution can only be found when some kind of regularization device is included in the inversion process. Inclusion of regularization in the calibration process can be implicit, for example through the use of zones of constant parameter value, or explicit, for example through solution of a constrained minimization problem in which parameters are made to respect preferred values, or preferred relationships, to the degree necessary for a unique solution to be obtained. The "cost of uniqueness" is this: no matter which regularization methodology is employed, the inevitable consequence of its use is a loss of detail in the calibrated field. This, in turn, can lead to erroneous predictions made by a model that is ostensibly "well calibrated". Information made available as a by-product of the regularized inversion process allows the reasons for this loss of detail to be better understood. In particular, it is easily demonstrated that the estimated value for an hydraulic property at any point within a model domain is, in fact, a weighted average of the true hydraulic property over a much larger area. This averaging process causes loss of resolution in the estimated field. Where hydraulic conductivity is the hydraulic property being estimated, high averaging weights exist in areas that are strategically disposed with respect to measurement wells, while other areas may contribute very little to the estimated hydraulic conductivity at any point within the model domain, this possibly making the detection of hydraulic conductivity anomalies in these latter areas almost impossible. A study of the post-calibration parameter field covariance matrix allows further insights into the loss of system detail incurred through the calibration process to be gained. A comparison of pre- and post-calibration

  7. A Microdrop Generator for the Calibration of a Water Vapor Isotope Ratio Spectrometer

    NARCIS (Netherlands)

    Iannone, Rosario Q.; Romanini, Daniele; Kassi, Samir; Meijer, Harro A. J.; Kerstel, Erik R. Th.

    A microdrop generator is described that produces water vapor with a known isotopic composition and volume mixing ratio for the calibration of a near-infrared diode laser water isotope ratio spectrometer. The spectrometer is designed to measure in situ the water vapor deuterium and oxygen ((17)O and

  8. Self-Calibration and Optimal Response in Intelligent Sensors Design Based on Artificial Neural Networks

    Directory of Open Access Journals (Sweden)

    Gilberto Bojorquez

    2007-08-01

    Full Text Available The development of smart sensors involves the design of reconfigurable systemscapable of working with different input sensors. Reconfigurable systems ideally shouldspend the least possible amount of time in their calibration. An autocalibration algorithmfor intelligent sensors should be able to fix major problems such as offset, variation of gainand lack of linearity, as accurately as possible. This paper describes a new autocalibrationmethodology for nonlinear intelligent sensors based on artificial neural networks, ANN.The methodology involves analysis of several network topologies and training algorithms.The proposed method was compared against the piecewise and polynomial linearizationmethods. Method comparison was achieved using different number of calibration points,and several nonlinear levels of the input signal. This paper also shows that the proposedmethod turned out to have a better overall accuracy than the other two methods. Besides,experimentation results and analysis of the complete study, the paper describes theimplementation of the ANN in a microcontroller unit, MCU. In order to illustrate themethod capability to build autocalibration and reconfigurable systems, a temperaturemeasurement system was designed and tested. The proposed method is an improvement over the classic autocalibration methodologies, because it impacts on the design process of intelligent sensors, autocalibration methodologies and their associated factors, like time and cost.

  9. Validation of a densimeter calibration procedure for a secondary calibration laboratory

    International Nuclear Information System (INIS)

    Alpizar Herrera, Juan Carlos

    2014-01-01

    A survey was conducted to quantify the need for calibration of a density measurement instrument at the research units at the Sede Rodrigo Facio of the Universidad de Costa Rica. A calibration procedure was documented for the instrument that presented the highest demand in the survey by the calibration service. A study of INTE-ISO/IEC 17025: 2005 and specifically in section 5.4 of this standard was done, to document the procedure for calibrating densimeters. Densimeter calibration procedures and standards were sought from different national and international sources. The method of hydrostatic weighing or Cuckow method was the basis of the defined procedure. Documenting the calibration procedure and creating other documents was performed for data acquisition log, intermediate calculation log and calibration certificate copy. A veracity test was performed using as reference laboratory a laboratory of calibration secondary national as part of the validation process of the documented procedure. The results of the E_n statistic of 0.41; 0.34 and 0.46 for the calibration points 90%, 50% and 10% were obtained for the densimeter scale respectively. A reproducibility analysis of the method was performed with satisfactory results. Different suppliers were contacted to estimate the economic costs of the equipment and materials, needed to develop the documented method of densimeter calibration. The acquisition of an analytical balance was recommended, instead of a precision scale, in order to improve the results obtained with the documented method [es

  10. A finite volume solver for three dimensional debris flow simulations based on a single calibration parameter

    Science.gov (United States)

    von Boetticher, Albrecht; Turowski, Jens M.; McArdell, Brian; Rickenmann, Dieter

    2016-04-01

    Debris flows are frequent natural hazards that cause massive damage. A wide range of debris flow models try to cover the complex flow behavior that arises from the inhomogeneous material mixture of water with clay, silt, sand, and gravel. The energy dissipation between moving grains depends on grain collisions and tangential friction, and the viscosity of the interstitial fine material suspension depends on the shear gradient. Thus a rheology description needs to be sensitive to the local pressure and shear rate, making the three-dimensional flow structure a key issue for flows in complex terrain. Furthermore, the momentum exchange between the granular and fluid phases should account for the presence of larger particles. We model the fine material suspension with a Herschel-Bulkley rheology law, and represent the gravel with the Coulomb-viscoplastic rheology of Domnik & Pudasaini (Domnik et al. 2013). Both composites are described by two phases that can mix; a third phase accounting for the air is kept separate to account for the free surface. The fluid dynamics are solved in three dimensions using the finite volume open-source code OpenFOAM. Computational costs are kept reasonable by using the Volume of Fluid method to solve only one phase-averaged system of Navier-Stokes equations. The Herschel-Bulkley parameters are modeled as a function of water content, volumetric solid concentration of the mixture, clay content and its mineral composition (Coussot et al. 1989, Yu et al. 2013). The gravel phase properties needed for the Coulomb-viscoplastic rheology are defined by the angle of repose of the gravel. In addition to this basic setup, larger grains and the corresponding grain collisions can be introduced by a coupled Lagrangian particle simulation. Based on the local Savage number a diffusive term in the gravel phase can activate phase separation. The resulting model can reproduce the sensitivity of the debris flow to water content and channel bed roughness, as

  11. 'Dip-sticks' calibration handles self-attenuation and coincidence effects in large-volume gamma-ray spectrometry

    CERN Document Server

    Wolterbeek, H T

    2000-01-01

    Routine gamma-spectrometric analyses of samples with low-level activities (e.g. food, water, environmental and industrial samples) are often performed in large samples, placed close to the detector. In these geometries, detection sensitivity is improved but large errors are introduced due to self-attenuation and coincidence summing. Current approaches to these problems comprise computational methods and spiked standard materials. However, the first are often regarded as too complex for practical routine use, the latter never fully match real samples. In the present study, we introduce a dip-sticks calibration as a fast and easy practical solution to this quantification problem in a routine analytical setting. In the proposed set-up, calibrations are performed within the sample itself, thus making it a broadly accessible matching-reference approach, which is principally usable for all sample matrices.

  12. Calibration of acoustic sensors in ice using the reciprocity method

    Energy Technology Data Exchange (ETDEWEB)

    Meures, Thomas; Bissok, Martin; Laihem, Karim; Paul, Larissa; Wiebusch, Christopher; Zierke, Simon [III. Physikalisches Institut, RWTH Aachen (Germany); Semburg, Benjamin [Bergische Universitaet Wuppertal (Germany). Fachbereich C

    2010-07-01

    Within the IceCube experiment at the South Pole an R and D program investigates new ways of ultra high energy neutrino detection. In particular when aiming for detector volumes of the order of 100 km{sup 3} acoustic or radio detectors are promising approaches. The acoustic detection method relies on the thermo-acoustic effect occurring when high energetic particles interact and deposit heat within a detection medium. This effect is investigated in the Aachen Acoustic Laboratory (AAL). The high energy particle interaction is simulated by a powerful pulsed Nd:YAG LASER shooting into a 3m{sup 3} tank of clear ice (or water). Eighteen acoustic sensors are situated on three rings in different depths and record the generated signals. These sensors serve as reference for later measurements of other devices. The reciprocity method, used for the absolute calibration of these sensors, is independent of an absolutely calibrated reference. This method and its application to the calibration of the AAL sensors are presented and first results are shown.

  13. LiDAR-IMU Time Delay Calibration Based on Iterative Closest Point and Iterated Sigma Point Kalman Filter.

    Science.gov (United States)

    Liu, Wanli

    2017-03-08

    The time delay calibration between Light Detection and Ranging (LiDAR) and Inertial Measurement Units (IMUs) is an essential prerequisite for its applications. However, the correspondences between LiDAR and IMU measurements are usually unknown, and thus cannot be computed directly for the time delay calibration. In order to solve the problem of LiDAR-IMU time delay calibration, this paper presents a fusion method based on iterative closest point (ICP) and iterated sigma point Kalman filter (ISPKF), which combines the advantages of ICP and ISPKF. The ICP algorithm can precisely determine the unknown transformation between LiDAR-IMU; and the ISPKF algorithm can optimally estimate the time delay calibration parameters. First of all, the coordinate transformation from the LiDAR frame to the IMU frame is realized. Second, the measurement model and time delay error model of LiDAR and IMU are established. Third, the methodology of the ICP and ISPKF procedure is presented for LiDAR-IMU time delay calibration. Experimental results are presented that validate the proposed method and demonstrate the time delay error can be accurately calibrated.

  14. Calibration of DEM parameters on shear test experiments using Kriging method

    Science.gov (United States)

    Xavier, Bednarek; Sylvain, Martin; Abibatou, Ndiaye; Véronique, Peres; Olivier, Bonnefoy

    2017-06-01

    Calibration of powder mixing simulation using Discrete-Element-Method is still an issue. Achieving good agreement with experimental results is difficult because time-efficient use of DEM involves strong assumptions. This work presents a methodology to calibrate DEM parameters using Efficient Global Optimization (EGO) algorithm based on Kriging interpolation method. Classical shear test experiments are used as calibration experiments. The calibration is made on two parameters - Young modulus and friction coefficient. The determination of the minimal number of grains that has to be used is a critical step. Simulations of a too small amount of grains would indeed not represent the realistic behavior of powder when using huge amout of grains will be strongly time consuming. The optimization goal is the minimization of the objective function which is the distance between simulated and measured behaviors. The EGO algorithm uses the maximization of the Expected Improvement criterion to find next point that has to be simulated. This stochastic criterion handles with the two interpolations made by the Kriging method : prediction of the objective function and estimation of the error made. It is thus able to quantify the improvement in the minimization that new simulations at specified DEM parameters would lead to.

  15. 37Ar based neutron source for calibration of the iodine solar neutrino detector

    International Nuclear Information System (INIS)

    Abdurashitov, D.N.; Gavrin, V.N.; Mirmov, I.N.; Veretenkin, E.P.; Yants, V.Eh.; Cleveland, B.T.; Davis, R. Jr.; Lande, K.; Wildenhain, P.; Khomyakov, Yu.S.

    2001-01-01

    The methodology of the creation of a compact neutrino source based on the 37 Ar isotope as well as the technique of calibration of an iodine detector of solar neutrinos is described. An important overall expected result is the creation of a prototype of the source with the intensity up to 400 kCi, delivery of this source to the Baksan neutrino observatory and the test calibration of the single module of the iodine detector. Simulation shows that at least 45-70 127 Xe atoms will be detected in the irradiation of ∼40 tons of methylene iodide by the source leading to ∼19% of the error on the measured production rate. This result should be considered as a test of the developed technology and will verify overall technical readiness for the creation of a full scale neutrino source and the full scale calibration of the iodine detector

  16. DART II documentation. Volume III. Appendices

    Energy Technology Data Exchange (ETDEWEB)

    1979-10-01

    The DART II is a remote, interactive, microprocessor-based data acquistion system suitable for use with air monitors. This volume of DART II documentation contains the following appendixes: adjustment and calibration procedures; mother board signature list; schematic diagrams; device specification sheets; ROM program listing; 6800 microprocessor instruction list, octal listing; and cable lists. (RWR)

  17. REDUCED GANGLION CELL VOLUME ON OPTICAL COHERENCE TOMOGRAPHY IN PATIENTS WITH GEOGRAPHIC ATROPHY.

    Science.gov (United States)

    Ramkumar, Hema L; Nguyen, Brian; Bartsch, Dirk-Uwe; Saunders, Luke J; Muftuoglu, Ilkay Kilic; You, Qisheng; Freeman, William R

    2017-11-07

    Geographic atrophy (GA) is the sequelae of macular degeneration. Automated inner retinal analysis using optical coherence tomography is flawed because segmentation software is calibrated for normal eyes. The purpose of this study is to determine whether ganglion cell layer (GCL) volume is reduced in GA using manual analysis. Nineteen eyes with subfoveal GA and 22 controls were selected for morphometric analyses. Heidelberg scanning laser ophthalmoscope optical coherence tomography images of the optic nerve and macula were obtained, and the Viewing Module was used to manually calibrate retinal layer segmentation. Retinal layer volumes in the central 3-mm and surrounding 6-mm diameter were measured. Linear mixed models were used for statistics. The GCL volume in the central 3 mm of the macula is less (P = 0.003), and the retinal nerve fiber layer volume is more (P = 0.02) in patients with GA when compared with controls. Ganglion cell layer volume positively correlated with outer nuclear layer volume (P = 0.020). The patients with geographic atrophy have a small significant loss of the GCL. Ganglion cell death may precede axonal loss, and increased macular retinal nerve fiber layer volumes are not indicative of GCL volume. Residual ganglion cell stimulation by interneurons may enable vision in patients with GA.

  18. Methodology for Multileaf Collimator Quality Assurance in clinical conditions

    International Nuclear Information System (INIS)

    Diaz M, R. M.; Rodriguez Z, M.; Juarez D, A.; Romero R, R.

    2013-01-01

    Multileaf Collimators (MLCs) have become an important technological advance as part of clinical linear accelerators (linacs) for radiotherapy. Treatment planning and delivery were substantially modified after these devices. However, it was needed to develop Quality Assurance (QA) methodologies related to the performance of these developments. The most common methods for QA of MLC are made in basic conditions that hardly cover all possible difficulties in clinical practice. Diaz et. el. developed a methodology based upon volumetric detectors bidimensional arrays that can be extended to more demanding situations. In this work, the Auril methodology of Diaz et. al. was implemented to the irradiation with the linac gantry in horizontal position. A mathematical procedure was developed to ease the dosimetric centering of the device with the Auril centering tool. System calibration was made as in the typical Auril methodology. Patterns with leaf misplacements in known positions were irradiated. the method allowed the detection of leafs' misplacements with a minimum number of false positives. We concluded that Auril methodology can be applied in clinical conditions. (Author)

  19. Role of calibration, validation, and relevance in multi-level uncertainty integration

    International Nuclear Information System (INIS)

    Li, Chenzhao; Mahadevan, Sankaran

    2016-01-01

    Calibration of model parameters is an essential step in predicting the response of a complicated system, but the lack of data at the system level makes it impossible to conduct this quantification directly. In such a situation, system model parameters are estimated using tests at lower levels of complexity which share the same model parameters with the system. For such a multi-level problem, this paper proposes a methodology to quantify the uncertainty in the system level prediction by integrating calibration, validation and sensitivity analysis at different levels. The proposed approach considers the validity of the models used for parameter estimation at lower levels, as well as the relevance at the lower level to the prediction at the system level. The model validity is evaluated using a model reliability metric, and models with multivariate output are considered. The relevance is quantified by comparing Sobol indices at the lower level and system level, thus measuring the extent to which a lower level test represents the characteristics of the system so that the calibration results can be reliably used in the system level. Finally the results of calibration, validation and relevance analysis are integrated in a roll-up method to predict the system output. - Highlights: • Relevance analysis to quantify the closeness of two models. • Stochastic model reliability metric to integrate multiple validation experiments. • Extend the model reliability metric to deal with multivariate output. • Roll-up formula to integrate calibration, validation, and relevance.

  20. A bronchoscopic navigation system using bronchoscope center calibration for accurate registration of electromagnetic tracker and CT volume without markers

    Energy Technology Data Exchange (ETDEWEB)

    Luo, Xiongbiao, E-mail: xiongbiao.luo@gmail.com [Robarts Research Institute, Western University, London, Ontario N6A 5K8 (Canada)

    2014-06-15

    Purpose: Various bronchoscopic navigation systems are developed for diagnosis, staging, and treatment of lung and bronchus cancers. To construct electromagnetically navigated bronchoscopy systems, registration of preoperative images and an electromagnetic tracker must be performed. This paper proposes a new marker-free registration method, which uses the centerlines of the bronchial tree and the center of a bronchoscope tip where an electromagnetic sensor is attached, to align preoperative images and electromagnetic tracker systems. Methods: The chest computed tomography (CT) volume (preoperative images) was segmented to extract the bronchial centerlines. An electromagnetic sensor was fixed at the bronchoscope tip surface. A model was designed and printed using a 3D printer to calibrate the relationship between the fixed sensor and the bronchoscope tip center. For each sensor measurement that includes sensor position and orientation information, its corresponding bronchoscope tip center position was calculated. By minimizing the distance between each bronchoscope tip center position and the bronchial centerlines, the spatial alignment of the electromagnetic tracker system and the CT volume was determined. After obtaining the spatial alignment, an electromagnetic navigation bronchoscopy system was established to real-timely track or locate a bronchoscope inside the bronchial tree during bronchoscopic examinations. Results: The electromagnetic navigation bronchoscopy system was validated on a dynamic bronchial phantom that can simulate respiratory motion with a breath rate range of 0–10 min{sup −1}. The fiducial and target registration errors of this navigation system were evaluated. The average fiducial registration error was reduced from 8.7 to 6.6 mm. The average target registration error, which indicates all tracked or navigated bronchoscope position accuracy, was much reduced from 6.8 to 4.5 mm compared to previous registration methods. Conclusions: An

  1. A bronchoscopic navigation system using bronchoscope center calibration for accurate registration of electromagnetic tracker and CT volume without markers

    International Nuclear Information System (INIS)

    Luo, Xiongbiao

    2014-01-01

    Purpose: Various bronchoscopic navigation systems are developed for diagnosis, staging, and treatment of lung and bronchus cancers. To construct electromagnetically navigated bronchoscopy systems, registration of preoperative images and an electromagnetic tracker must be performed. This paper proposes a new marker-free registration method, which uses the centerlines of the bronchial tree and the center of a bronchoscope tip where an electromagnetic sensor is attached, to align preoperative images and electromagnetic tracker systems. Methods: The chest computed tomography (CT) volume (preoperative images) was segmented to extract the bronchial centerlines. An electromagnetic sensor was fixed at the bronchoscope tip surface. A model was designed and printed using a 3D printer to calibrate the relationship between the fixed sensor and the bronchoscope tip center. For each sensor measurement that includes sensor position and orientation information, its corresponding bronchoscope tip center position was calculated. By minimizing the distance between each bronchoscope tip center position and the bronchial centerlines, the spatial alignment of the electromagnetic tracker system and the CT volume was determined. After obtaining the spatial alignment, an electromagnetic navigation bronchoscopy system was established to real-timely track or locate a bronchoscope inside the bronchial tree during bronchoscopic examinations. Results: The electromagnetic navigation bronchoscopy system was validated on a dynamic bronchial phantom that can simulate respiratory motion with a breath rate range of 0–10 min −1 . The fiducial and target registration errors of this navigation system were evaluated. The average fiducial registration error was reduced from 8.7 to 6.6 mm. The average target registration error, which indicates all tracked or navigated bronchoscope position accuracy, was much reduced from 6.8 to 4.5 mm compared to previous registration methods. Conclusions: An

  2. Determination of fragrance content in perfume by Raman spectroscopy and multivariate calibration.

    Science.gov (United States)

    Godinho, Robson B; Santos, Mauricio C; Poppi, Ronei J

    2016-03-15

    An alternative methodology is herein proposed for determination of fragrance content in perfumes and their classification according to the guidelines established by fine perfume manufacturers. The methodology is based on Raman spectroscopy associated with multivariate calibration, allowing the determination of fragrance content in a fast, nondestructive, and sustainable manner. The results were considered consistent with the conventional method, whose standard error of prediction values was lower than the 1.0%. This result indicates that the proposed technology is a feasible analytical tool for determination of the fragrance content in a hydro-alcoholic solution for use in manufacturing, quality control and regulatory agencies. Copyright © 2015 Elsevier B.V. All rights reserved.

  3. ORNL calibrations facility

    International Nuclear Information System (INIS)

    Berger, C.D.; Gupton, E.D.; Lane, B.H.; Miller, J.H.; Nichols, S.W.

    1982-08-01

    The ORNL Calibrations Facility is operated by the Instrumentation Group of the Industrial Safety and Applied Health Physics Division. Its primary purpose is to maintain radiation calibration standards for calibration of ORNL health physics instruments and personnel dosimeters. This report includes a discussion of the radioactive sources and ancillary equipment in use and a step-by-step procedure for calibration of those survey instruments and personnel dosimeters in routine use at ORNL

  4. SCIAMACHY Level 1 data: calibration concept and in-flight calibration

    Science.gov (United States)

    Lichtenberg, G.; Kleipool, Q.; Krijger, J. M.; van Soest, G.; van Hees, R.; Tilstra, L. G.; Acarreta, J. R.; Aben, I.; Ahlers, B.; Bovensmann, H.; Chance, K.; Gloudemans, A. M. S.; Hoogeveen, R. W. M.; Jongma, R. T. N.; Noël, S.; Piters, A.; Schrijver, H.; Schrijvers, C.; Sioris, C. E.; Skupin, J.; Slijkhuis, S.; Stammes, P.; Wuttke, M.

    2006-11-01

    The calibration of SCIAMACHY was thoroughly checked since the instrument was launched on-board ENVISAT in February 2002. While SCIAMACHY's functional performance is excellent since launch, a number of technical difficulties have appeared, that required adjustments to the calibration. The problems can be separated into three types: (1) Those caused by the instrument and/or platform environment. Among these are the high water content in the satellite structure and/or MLI layer. This results in the deposition of ice on the detectors in channels 7 and 8 which seriously affects the retrievals in the IR, mostly because of the continuous change of the slit function caused by scattering of the light through the ice layer. Additionally a light leak in channel 7 severely hampers any retrieval from this channel. (2) Problems due to errors in the on-ground calibration and/or data processing affecting for example the radiometric calibration. A new approach based on a mixture of on-ground and in-flight data is shortly described here. (3) Problems caused by principal limitations of the calibration concept, e.g. the possible appearance of spectral structures after the polarisation correction due to unavoidable errors in the determination of atmospheric polarisation. In this paper we give a complete overview of the calibration and problems that still have to be solved. We will also give an indication of the effect of calibration problems on retrievals where possible. Since the operational processing chain is currently being updated and no newly processed data are available at this point in time, for some calibration issues only a rough estimate of the effect on Level 2 products can be given. However, it is the intention of this paper to serve as a future reference for detailed studies into specific calibration issues.

  5. Using genetic algorithm and TOPSIS for Xinanjiang model calibration with a single procedure

    Science.gov (United States)

    Cheng, Chun-Tian; Zhao, Ming-Yan; Chau, K. W.; Wu, Xin-Yu

    2006-01-01

    Genetic Algorithm (GA) is globally oriented in searching and thus useful in optimizing multiobjective problems, especially where the objective functions are ill-defined. Conceptual rainfall-runoff models that aim at predicting streamflow from the knowledge of precipitation over a catchment have become a basic tool for flood forecasting. The parameter calibration of a conceptual model usually involves the multiple criteria for judging the performances of observed data. However, it is often difficult to derive all objective functions for the parameter calibration problem of a conceptual model. Thus, a new method to the multiple criteria parameter calibration problem, which combines GA with TOPSIS (technique for order performance by similarity to ideal solution) for Xinanjiang model, is presented. This study is an immediate further development of authors' previous research (Cheng, C.T., Ou, C.P., Chau, K.W., 2002. Combining a fuzzy optimal model with a genetic algorithm to solve multi-objective rainfall-runoff model calibration. Journal of Hydrology, 268, 72-86), whose obvious disadvantages are to split the whole procedure into two parts and to become difficult to integrally grasp the best behaviors of model during the calibration procedure. The current method integrates the two parts of Xinanjiang rainfall-runoff model calibration together, simplifying the procedures of model calibration and validation and easily demonstrated the intrinsic phenomenon of observed data in integrity. Comparison of results with two-step procedure shows that the current methodology gives similar results to the previous method, is also feasible and robust, but simpler and easier to apply in practice.

  6. X-ray and gamma-ray standards for detector calibration

    International Nuclear Information System (INIS)

    1991-09-01

    The IAEA established a Co-ordinated Research Programme (CRP) on the Measurements and Evaluation of X- and Gamma-Ray Standards for Detector Efficiency Calibration in 1986 with the aim of alleviating the generation of such discrepancies. Within the framework of this CRP, representatives of nine research groups from six Member States and one international organization performed a number of precise measurements and systematic in-depth evaluations of the required decay data. They have also contributed to the development of evaluation methodology and measurement techniques, and stimulated a number of such studies at laboratories not directly involved in the IAEA project. The results of the work of the CRP, which was finished in 1990, are presented in this report. Recommended values of half-lives and photon emission probabilities are given for a carefully selected set of radionuclides that are suitable for detector efficiency calibration (X-rays from 5 to 90 keV and gamma-rays from 30 to about 3000 keV). Detector efficiency calibration for higher gamma-ray energies (up to 14 MeV) is also considered. The evaluation procedures used to obtain the recommended values and their estimated uncertainties are reported, and a summary of the remaining discrepancies is given. Refs and tabs

  7. Metrological tests of a 200 L calibration source for HPGE detector systems for assay of radioactive waste drums.

    Science.gov (United States)

    Boshkova, T; Mitev, K

    2016-03-01

    In this work we present test procedures, approval criteria and results from two metrological inspections of a certified large volume (152)Eu source (drum about 200L) intended for calibration of HPGe gamma assay systems used for activity measurement of radioactive waste drums. The aim of the inspections was to prove the stability of the calibration source during its working life. The large volume source was designed and produced in 2007. It consists of 448 identical sealed radioactive sources (modules) apportioned in 32 transparent plastic tubes which were placed in a wooden matrix which filled the drum. During the inspections the modules were subjected to tests for verification of their certified characteristics. The results show a perfect compliance with the NIST basic guidelines for the properties of a radioactive certified reference material (CRM) and demonstrate the stability of the large volume CRM-drum after 7 years of operation. Copyright © 2015 Elsevier Ltd. All rights reserved.

  8. Technological considerations in emergency instrumentation preparedness. Phase II-D. Evaluation testing and calibration methodology for emergency radiological instrumentation

    International Nuclear Information System (INIS)

    Bramson, P.E.; Andersen, B.V.; Fleming, D.M.; Kathren, R.L.; Mulhern, O.R.; Newton, C.E.; Oscarson, E.E.; Selby, J.M.

    1976-09-01

    In response to recommendations from the Advisory Committee on Reactor Safeguards, the Division of Operational Safety, U.S. ERDA has contracted with Battelle, Pacific Northwest Laboratories to survey the adequacy of existing instrumentation at nuclear fuel cycle facilities to meet emergency requirements and to develop technical criteria for instrumentation systems to be used in assessment of environmental conditions following plant emergencies. This report, the fifth in a series, provides: (1) calibration methods to assure the quality of radiological measurements and (2) testing procedures for determining whether an emergency radiological instrument meets the performance specifications. Three previous reports in this series identified the emergency instrumentation needs for power reactors, mixed oxide fuel plants, and fuel reprocessing facilities. Each of these three reports contains a Section VI, which sets forth applicable radiological instrument performance criteria and calibration requirements. Testing and calibration procedures in this report have been formatted in two parts: IV and V, each divided into three subsections: (1) Power Reactors, (2) Mixed Oxide Fuel Plants, and (3) Fuel Reprocessing Facilities. The three performance criteria subsections directly coincide with the performance criteria sections of the previous reports. These performance criteria sections have been reproduced in this report as Part III with references of ''required action'' added

  9. Oral cancer calibration and diagnosis among professionals from the public health in São Paulo, Brazil.

    Science.gov (United States)

    Alves, José Carlos; da Silva, Renato Pereira; Cortellazzi, Karine Laura; Vazquez, Fabiana de Lima; Marques, Regina Auxiliadora de Amorim; Pereira, Antonio Carlos; Meneghim, Marcelo de Castro; Mialhe, Fábio Luiz

    2013-01-01

    Oral cancer is a public health problem responsible for 13% of deaths worldwide in 2008 and screening programs can be useful to detect individuals more vulnerable to the disease, improving its prognosis. The aim of the present study was to evaluate oral cancer calibration (in lux and in vivo methodologies) and diagnosis processes performed by dental surgeons (DSs) of the public health system in São Paulo, Brazil. Thirty-three oral cancer photographs were examined during in lux calibration, while 560 individuals were examined during in vivo calibration. Oral conditions were coded as "0 - sound tissues", "1 - buccal lesions without malignant potential" and "2 - buccal lesions with malignant potential". The final sample for oral cancer screening was composed of 336 individuals, age-range 40 years or older. Kappa values for interexaminer agreement were 0.67 and 0.45 for in lux and in vivo respectively. The accuracy of both methodologies was over 80%. Oral cancer screening revealed 48 healthy individuals, 273 oral lesions coded as "1" and 12 oral lesions coded as "2". In spite of the low reproducibility, the validity of the visual examination in oral cancer screening was satisfactory, showing its importance as part of preventive oral cancer programs and public health system campaigns.

  10. Assessment of a dynamic reference material for calibration of full-field measurement systems

    Science.gov (United States)

    Hack, Erwin; Feligiotti, Mara; Davighi, Andrea; Whelan, Maurice; Wang, Weizhuo V.; Patterson, Eann A.

    2012-10-01

    For holography and speckle interferometry the calibration of the sensitivity is a must, because illumination and observation directions vary across the field of view. A numerical estimate or a static calibration using rigid body motions is standard, and reference materials exist for static strain calibration. Recently, reference materials for the dynamic calibration of optical instruments of displacement and strain measurement were designed and prototypes were manufactured in the European FP7 project ADVISE. We review the properties of the reference material and the concept of traceability for the field of displacement values by using a calibrated single point transducer. The mode shape is assessed using out-of-plane DSPI, Finite Element Analysis as well as analytic solutions of the plate vibration. We present measurements using stroboscopic DSPI on the reference material under acoustic excitation and compare the measured mode shapes to the ones predicted by FE analysis. We apply different comparison methodologies based on point-by-point deviations and on decomposition of the mode shapes into a set of orthogonal basis functions. The latter method is well suited to assess stability and reproducibility of a mode shape. Finally, the deviations are used to estimate the reference material uncertainty which is an essential parameter for determining the calibration uncertainty. Uncertainty contributions of the DSPI set-up are taken into account. To conclude, the application area and limitations of the reference material are discussed.

  11. Uncertainty determination in a custody transfer operation from vertical cylinder storage tanks; Determinacao da incerteza do volume transferido em tanques cilindricos verticais

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Elcio C.; Ferreira, Ana Luisa A.S. [TRANSPETRO - PETROBRAS Transporte S.A., Rio de Janeiro, RJ (Brazil); Orlando, Alcir F.; Val, Luiz G. do [Pontificia Univ. Catolica do Rio de Janeiro, RJ (Brazil)

    2004-07-01

    The INMETRO/ANP 1 regulation (2000), presents rules to be followed for measuring and calibrating cylindrical vertical oil storage tanks in Brazil, according to ISO 7507-1 (1993) standard. A methodology for estimating the uncertainty (95,45 % confidence level) of the volume in a custody transfer process was developed, based on ISO GUM (1998) standard. The strapping method was selected for this study, because it has been used as a standard procedure by INMETRO. In this study, the same uncertainty values, as suggested by the standard, were used to estimate the uncertainty of the liquid volume in the tank. This study showed that the uncertainty of the transferred liquid volume from the tank varies from 0,2% to 0,4%, being smaller for larger volumes, which is thus the recommended application. The uncertainty of the ring height measurement is the largest contribution to the volume measurement uncertainty, and, thus, must be accurately measured. The tank internal diameter uncertainty is a small contribution to it. This paper calculates the uncertainty of liquid volume transferred from the tank by three methods, namely, this paper's, ISO 7507- 1's and INMETRO's, and shows that the most important contribution to the measurement uncertainty is the density measurement uncertainty, which must be accurately measured, at least, to within {+-} 0,0005, if the volume uncertainty is to remain in the 0,5 % to 1 % range. (author)

  12. MoDOT pavement preservation research program volume VII, re-calibration of triggers and performance models.

    Science.gov (United States)

    2015-10-01

    The objective of this task is to develop the concept and framework for a procedure to routinely create, re-calibrate, and update the : Trigger Tables and Performance Models. The scope of work for Task 6 includes a limited review of the recent pavemen...

  13. Spitzer/JWST Cross Calibration: IRAC Observations of Potential Calibrators for JWST

    Science.gov (United States)

    Carey, Sean J.; Gordon, Karl D.; Lowrance, Patrick; Ingalls, James G.; Glaccum, William J.; Grillmair, Carl J.; E Krick, Jessica; Laine, Seppo J.; Fazio, Giovanni G.; Hora, Joseph L.; Bohlin, Ralph

    2017-06-01

    We present observations at 3.6 and 4.5 microns using IRAC on the Spitzer Space Telescope of a set of main sequence A stars and white dwarfs that are potential calibrators across the JWST instrument suite. The stars range from brightnesses of 4.4 to 15 mag in K band. The calibration observations use a similar redundancy to the observing strategy for the IRAC primary calibrators (Reach et al. 2005) and the photometry is obtained using identical methods and instrumental photometric corrections as those applied to the IRAC primary calibrators (Carey et al. 2009). The resulting photometry is then compared to the predictions based on spectra from the CALSPEC Calibration Database (http://www.stsci.edu/hst/observatory/crds/calspec.html) and the IRAC bandpasses. These observations are part of an ongoing collaboration between IPAC and STScI investigating absolute calibration in the infrared.

  14. Piezoelectric trace vapor calibrator

    International Nuclear Information System (INIS)

    Verkouteren, R. Michael; Gillen, Greg; Taylor, David W.

    2006-01-01

    The design and performance of a vapor generator for calibration and testing of trace chemical sensors are described. The device utilizes piezoelectric ink-jet nozzles to dispense and vaporize precisely known amounts of analyte solutions as monodisperse droplets onto a hot ceramic surface, where the generated vapors are mixed with air before exiting the device. Injected droplets are monitored by microscope with strobed illumination, and the reproducibility of droplet volumes is optimized by adjustment of piezoelectric wave form parameters. Complete vaporization of the droplets occurs only across a 10 deg. C window within the transition boiling regime of the solvent, and the minimum and maximum rates of trace analyte that may be injected and evaporated are determined by thermodynamic principles and empirical observations of droplet formation and stability. By varying solution concentrations, droplet injection rates, air flow, and the number of active nozzles, the system is designed to deliver--on demand--continuous vapor concentrations across more than six orders of magnitude (nominally 290 fg/l to 1.05 μg/l). Vapor pulses containing femtogram to microgram quantities of analyte may also be generated. Calibrated ranges of three explosive vapors at ng/l levels were generated by the device and directly measured by ion mobility spectrometry (IMS). These data demonstrate expected linear trends within the limited working range of the IMS detector and also exhibit subtle nonlinear behavior from the IMS measurement process

  15. Calibrating mechanistic-empirical pavement performance models with an expert matrix

    Energy Technology Data Exchange (ETDEWEB)

    Tighe, S.; AlAssar, R.; Haas, R. [Waterloo Univ., ON (Canada). Dept. of Civil Engineering; Zhiwei, H. [Stantec Consulting Ltd., Cambridge, ON (Canada)

    2001-07-01

    Proper management of pavement infrastructure requires pavement performance modelling. For the past 20 years, the Ontario Ministry of Transportation has used the Ontario Pavement Analysis of Costs (OPAC) system for pavement design. Pavement needs, however, have changed substantially during that time. To address this need, a new research contract is underway to enhance the model and verify the predictions, particularly at extreme points such as low and high traffic volume pavement design. This initiative included a complete evaluation of the existing OPAC pavement design method, the construction of a new set of pavement performance prediction models, and the development of the flexible pavement design procedure that incorporates reliability analysis. The design was also expanded to include rigid pavement designs and modification of the existing life cycle cost analysis procedure which includes both the agency cost and road user cost. Performance prediction and life-cycle costs were developed based on several factors, including material properties, traffic loads and climate. Construction and maintenance schedules were also considered. The methodology for the calibration and validation of a mechanistic-empirical flexible pavement performance model was described. Mechanistic-empirical design methods combine theory based design such as calculated stresses, strains or deflections with empirical methods, where a measured response is associated with thickness and pavement performance. Elastic layer analysis was used to determine pavement response to determine the most effective design using cumulative Equivalent Single Axle Loads (ESALs), below grade type and layer thickness.The new mechanistic-empirical model separates the environment and traffic effects on performance. This makes it possible to quantify regional differences between Southern and Northern Ontario. In addition, roughness can be calculated in terms of the International Roughness Index or Riding comfort Index

  16. Conceptual study of calibration software for large scale input accountancy tank

    International Nuclear Information System (INIS)

    Uchikoshi, Seiji; Yasu, Kan-ichi; Watanabe, Yuichi; Matsuda, Yuji; Kawai, Akio; Tamura, Toshiyuki; Shimizu, Hidehiko.

    1996-01-01

    Demonstration experiments for large scale input accountancy tank are going to be under way by Nuclear Material Control Center. Development of calibration software for accountancy system with dip-tube manometer is an important task in the experiments. A conceptual study of the software has been carried out to construct high precision accountancy system. And, the study was based on ANSI N15.19-1989. Items of the study are overall configuration, correction method for influence of bubble formation, function model of calibration, and fitting method for calibration curve. Following remarks are the results of this study. 1) Overall configuration of the software was constructed. 2) It was shown by numerical solution, that the influence of bubble formation can be corrected using period of pressure wave. 3) Two function models of calibration for well capacity and for inner structure volume were prepared from tank design, and good fitness of the model for net capacity (balance of both models) was confirmed by fitting to designed shape of the tank. 4) The necessity of further consideration about both-variables-in-error-model and cumulative-error-model was recognized. We are going to develop a practical software on the basis of the results, and to verify it by the demonstration experiments. (author)

  17. Simultaneous calibration phantom commission and geometry calibration in cone beam CT

    Science.gov (United States)

    Xu, Yuan; Yang, Shuai; Ma, Jianhui; Li, Bin; Wu, Shuyu; Qi, Hongliang; Zhou, Linghong

    2017-09-01

    Geometry calibration is a vital step for describing the geometry of a cone beam computed tomography (CBCT) system and is a prerequisite for CBCT reconstruction. In current methods, calibration phantom commission and geometry calibration are divided into two independent tasks. Small errors in ball-bearing (BB) positioning in the phantom-making step will severely degrade the quality of phantom calibration. To solve this problem, we propose an integrated method to simultaneously realize geometry phantom commission and geometry calibration. Instead of assuming the accuracy of the geometry phantom, the integrated method considers BB centers in the phantom as an optimized parameter in the workflow. Specifically, an evaluation phantom and the corresponding evaluation contrast index are used to evaluate geometry artifacts for optimizing the BB coordinates in the geometry phantom. After utilizing particle swarm optimization, the CBCT geometry and BB coordinates in the geometry phantom are calibrated accurately and are then directly used for the next geometry calibration task in other CBCT systems. To evaluate the proposed method, both qualitative and quantitative studies were performed on simulated and realistic CBCT data. The spatial resolution of reconstructed images using dental CBCT can reach up to 15 line pair cm-1. The proposed method is also superior to the Wiesent method in experiments. This paper shows that the proposed method is attractive for simultaneous and accurate geometry phantom commission and geometry calibration.

  18. A data-calibrated distribution of deglacial chronologies for the North American ice complex from glaciological modeling

    Science.gov (United States)

    Tarasov, Lev; Dyke, Arthur S.; Neal, Radford M.; Peltier, W. R.

    2012-01-01

    Past deglacial ice sheet reconstructions have generally relied upon discipline-specific constraints with no attention given to the determination of objective confidence intervals. Reconstructions based on geophysical inversion of relative sea level (RSL) data have the advantage of large sets of proxy data but lack ice-mechanical constraints. Conversely, reconstructions based on dynamical ice sheet models are glaciologically self-consistent, but depend on poorly constrained climate forcings and sub-glacial processes. As an example of a much better constrained methodology that computes explicit error bars, we present a distribution of high-resolution glaciologically-self-consistent deglacial histories for the North American ice complex calibrated against a large set of RSL, marine limit, and geodetic data. The history is derived from ensemble-based analyses using the 3D MUN glacial systems model and a high-resolution ice-margin chronology derived from geological and geomorphological observations. Isostatic response is computed with the VM5a viscosity structure. Bayesian calibration of the model is carried out using Markov Chain Monte Carlo methods in combination with artificial neural networks trained to the model results. The calibration provides a posterior distribution for model parameters (and thereby modeled glacial histories) given the observational data sets that takes data uncertainty into account. Final ensemble results also account for fits between computed and observed strandlines and marine limits. Given the model (including choice of calibration parameters), input and constraint data sets, and VM5a earth rheology, we find the North American contribution to mwp1a was likely between 9.4 and 13.2 m eustatic over a 500 year interval. This is more than half of the total 16 to 26 m meltwater pulse over 500 to 700 years (with lower values being more probable) indicated by the Barbados coral record (Fairbanks, 1989; Peltier and Fairbanks, 2006) if one assumes a

  19. Calibration of NRSF2 Instrument at HFIR

    International Nuclear Information System (INIS)

    Tang, Fei; Hubbard, Camden R.

    2006-01-01

    the sample table, 2θ 0 offset, and response of the position sensitive detector; (2) Counting statistics - if the peak profile count is too low, then the peak position derived from fitting a profile and background will have larger error. Therefore, adequate counting statistics and well-defined peaks are always good for precise peak position determination; and (3) Sample - Large grain size materials make it difficult to get enough diffracting grains, contributing to the different profile. With a low number the peak becomes 'spot' and results in inaccuracy in peak position. Texture in the sample can change the effective elastic constants and also affect the peak intensity. Phase and composition inhomogeneity can make it difficult to determine an accurate stress-free d 0 for strain calculation. A partially buried gauge volume due to proximity to the sample surface or buried interface can also shift the peak position. The calibration method presented in this report will address the first two categories of difficulties listed above. The FWHM can be minimized for each sample d-spacing by adjusting the horizontal bending of the monochromator crystal. For the monochromator, the optimum FWHM lies between 70 and 110 degree. This range is selected in order to maintain an approximately equiaxed gauge volume and avoid significant increases in peak breadth for the detectors above and below the horizontal plane. To adequately calibrate the position sensitive detectors, 2θ 0 , and wavelength, a set of high purity reference powders were selected. Since the selected reference powders have define grain size is, the measurement errors from sample grain size and texture can be excluded, although there may still be micro-strain in the powders, which can broaden the reference peak. In this report, the calibration procedure for the NRSF2 instrument will be presented and calibration results for five monochromator settings from HFIR cycle 403 will be presented. The monochromator settings

  20. SPRT Calibration Uncertainties and Internal Quality Control at a Commercial SPRT Calibration Facility

    Science.gov (United States)

    Wiandt, T. J.

    2008-06-01

    The Hart Scientific Division of the Fluke Corporation operates two accredited standard platinum resistance thermometer (SPRT) calibration facilities, one at the Hart Scientific factory in Utah, USA, and the other at a service facility in Norwich, UK. The US facility is accredited through National Voluntary Laboratory Accreditation Program (NVLAP), and the UK facility is accredited through UKAS. Both provide SPRT calibrations using similar equipment and procedures, and at similar levels of uncertainty. These uncertainties are among the lowest available commercially. To achieve and maintain low uncertainties, it is required that the calibration procedures be thorough and optimized. However, to minimize customer downtime, it is also important that the instruments be calibrated in a timely manner and returned to the customer. Consequently, subjecting the instrument to repeated calibrations or extensive repeated measurements is not a viable approach. Additionally, these laboratories provide SPRT calibration services involving a wide variety of SPRT designs. These designs behave differently, yet predictably, when subjected to calibration measurements. To this end, an evaluation strategy involving both statistical process control and internal consistency measures is utilized to provide confidence in both the instrument calibration and the calibration process. This article describes the calibration facilities, procedure, uncertainty analysis, and internal quality assurance measures employed in the calibration of SPRTs. Data will be reviewed and generalities will be presented. Finally, challenges and considerations for future improvements will be discussed.

  1. Methodology for calibration of detector of Nal (TI) 3 'X 3' for measurements in vivo of patients with hyperthyroidism undergoing radioiodine therapy; Metodologia para calibracao de detector de Nal(TI) 3'X3' para medicoes in vivo em pacientes portadores de hipertireoidismo submetidos a radioiodoterapia

    Energy Technology Data Exchange (ETDEWEB)

    Carvalho, Carlaine B.; Lacerda, Isabelle V.B.; Oliveira, Mercia L.; Hazin, Clovis A.; Lima, Fabiana F., E-mail: carlaine.carvalho@gmail.com, E-mail: bellelacerda@hotmail.com, E-mail: mercial@cnen.gov.br, E-mail: chazin@cnen.gov.br, E-mail: fflima@cnen.gov.br [Centro Regional de Ciencias Nucleares do Nordeste (CRCN-Ne/CNEN-PE), Recife, PE (Brazil)

    2013-11-01

    The aim of this study is to establish the methodology for calibration of the detection system to be used in determining the therapeutic activity of {sup 131}I required to release the desired absorbed dose in the thyroid. This step is critical to the development of a protocol for individualized doses. The system consists of a detector of NaI (Tl ) 3 'x3' coupled to Genie 2000 software. The calibration sources of {sup 60}CO, {sup 137}Cs and {sup 133}Ba were used. Obtained straight calibration system, with {sup 60}CO and {sup 137}Cs sources. Subsequently, the detector was calibrated using a simulator -neck thyroid designed and produced by the IRD/CNEN with known standard solution containing 18.7 kBq {sup 133}Ba activity (in 12/09/24) evenly distributed. He was also calibrated with other thyroid - neck phantom Model 3108 manufactured by Searle Radigraphics Ind., containing a net source of {sup 131}I ( 7.7 MBq ). Five measurements of five minutes were realized for three different distances detector simulator, and the respective calculated calibration factors was performed to three. The values of the calibration factors found for the simulator manufactured by IRD and the Searle Radigraphics Ind. for the distances 20 , 25 and 30cm were 0.35, 0.24, 0.18, and 0.15, 0.11, 0.09, respectively. With the detection system properly calibrated and the calibration factors established, the technique is suitable for the evaluation of diagnostic activity of {sup 131}I incorporated by hyperthyroidism.

  2. Calibration and quality assurance for rounded leaf-end MLC systems

    International Nuclear Information System (INIS)

    Graves, Maria N.; Thompson, Antoinette V.; Martel, Mary K.; McShan, Daniel L.; Fraass, Benedick A.

    2001-01-01

    Multileaf collimator (MLC) systems are available on most commercial linear accelerators, and many of these MLC systems utilize a design with rounded leaf ends and linear motion of the leaves. In this kind of system, the agreement between the digital MLC position readouts and the light field or radiation field edges must be achieved with software, since the leaves do not move in a focused motion like that used for most collimator jaw systems. In this work we address a number of the calibration and quality assurance issues associated with the acceptance, commissioning, and routine clinical use of this type of MLC system. These issues are particularly important for MLCs used for various types of intensity modulated radiation therapy (IMRT) and small, conformal fields. For rounded leaf end MLCs, it is generally not possible to make both the light and radiation field edges agree with the digital readout, so differences between the two kinds of calibrations are illustrated in this work using one vendor's MLC system. It is increasingly critical that the MLC leaf calibration be very consistent with the radiation field edges, so in this work a methodology for performing accurate radiation field size calibration is discussed. A system external to the vendor's MLC control system is used to correct or handle limitations in the MLC control system. When such a system of corrections is utilized, it is found that the MLC radiation field size can be defined with an accuracy of approximately 0.3 mm, much more accurate than most vendor's specifications for MLC accuracy. Quality assurance testing for such a calibration correction system is also demonstrated

  3. Calibration of Super-Kamiokande using an electron LINAC The Super-Kamiokande Collaboration

    CERN Document Server

    Nakahata, M; Hayakawa, T

    1999-01-01

    In order to calibrate the Super-Kamiokande experiment for solar neutrino measurements, a linear accelerator (LINAC) for electrons was installed at the detector. LINAC data were taken at various positions in the detector volume, tracking the detector response in the variables relevant to solar neutrino analysis. In particular, the absolute energy scale is now known with less than 1% uncertainty.

  4. Calibration of a Noble Gas Mass Spectrometer with an Atmospheric Argon Standard (Invited)

    Science.gov (United States)

    Prasad, V.; Grove, M.

    2009-12-01

    Like other mass spectrometers, gas source instruments are very good at precisely measuring isotopic ratios but need to be calibrated with a standard to be accurate. The need for calibration arises due to the complicated ionization process which inefficiently and differentially creates ions from the various isotopes that make up the elemental gas. Calibration of the ionization process requires materials with well understood isotopic compositions as standards. Our project goal was to calibrate a noble gas (Noblesse) mass spectrometer with a purified air sample. Our sample obtained from Ocean Beach in San Francisco was under known temperature, pressure, volume, humidity. We corrected the pressure for humidity and used the ideal gas law to calculate the number of moles of argon gas. We then removed all active gasses using specialized equipment designed for this purpose at the United States Geological Survey. At the same time, we measured the volume ratios of various parts of the gas extraction line system associated with the Noblesse mass spectrometer. Using this data, we calculated how much Ar was transferred to the reservoir from the vacuum-sealed vial that contained the purified gas standard. Using similar measurements, we also calculated how much Ar was introduced into the extraction line from a pipette system and how much of this Ar was ultimately expanded into the Noblesse mass spectrometer. Based upon this information, it was possible to calibrate the argon sensitivity of the mass spectrometer. From a knowledge of the isotopic composition of air, it was also possible to characterize how ionized argon isotopes were fractionated during analysis. By repeatedly analyzing our standard we measured a 40Ar Sensitivity of 2.05 amps/bar and a 40Ar/36Ar ratio of 309.2 on the Faraday detector. In contrast, measurements carried out by ion counting using electron multipliers yield a value (296.8) which is much closer to the actual atmospheric 40Ar/36Ar value of 295.5.

  5. Estimation of surface area and pore volume of activated carbons by methylene blue and iodine numbers

    Directory of Open Access Journals (Sweden)

    Cleiton A. Nunes

    2011-01-01

    Full Text Available Data of methylene blue number and iodine number of activated carbons samples were calibrated against the respective surface area, micropore volume and total pore volume using multiple regression. The models obtained from the calibrations were used in predicting these physical properties of a test group of activated carbon samples produced from several raw materials. In all cases, the predicted values were in good agreement with the expected values. The method allows extracting more information from the methylene blue and iodine adsorption studies than normally obtained with this type of material.

  6. Mechanics of log calibration

    International Nuclear Information System (INIS)

    Waller, W.C.; Cram, M.E.; Hall, J.E.

    1975-01-01

    For any measurement to have meaning, it must be related to generally accepted standard units by a valid and specified system of comparison. To calibrate well-logging tools, sensing systems are designed which produce consistent and repeatable indications over the range for which the tool was intended. The basics of calibration theory, procedures, and calibration record presentations are reviewed. Calibrations for induction, electrical, radioactivity, and sonic logging tools will be discussed. The authors' intent is to provide an understanding of the sources of errors, of the way errors are minimized in the calibration process, and of the significance of changes in recorded calibration data

  7. Nuclear Dynamics Consequence Analysis (NDCA) for the Disposal of Spent Nuclear Fuel in an Underground Geologic Repository--Volume 2: Methodology and Results

    International Nuclear Information System (INIS)

    Taylor, L.L.; Wilson, J.R.; Sanchez, L.C.; Aguilar, R.; Trellue, H.R.; Cochrane, K.; Rath, J.S.

    1998-01-01

    The US Department of Energy Office of Environmental Management's (DOE/EM's) National Spent Nuclear Fuel Program (NSNFP), through a collaboration between Sandia National Laboratories (SNL) and Idaho National Engineering and Environmental Laboratory (INEEL), is conducting a systematic Nuclear Dynamics Consequence Analysis (NDCA) of the disposal of SNFs in an underground geologic repository sited in unsaturated tuff. This analysis is intended to provide interim guidance to the DOE for the management of the SNF while they prepare for final compliance evaluation. This report presents results from a Nuclear Dynamics Consequence Analysis (NDCA) that examined the potential consequences and risks of criticality during the long-term disposal of spent nuclear fuel owned by DOE-EM. This analysis investigated the potential of post-closure criticality, the consequences of a criticality excursion, and the probability frequency for post-closure criticality. The results of the NDCA are intended to provide the DOE-EM with a technical basis for measuring risk which can be used for screening arguments to eliminate post-closure criticality FEPs (features, events and processes) from consideration in the compliance assessment because of either low probability or low consequences. This report is composed of an executive summary (Volume 1), the methodology and results of the NDCA (Volume 2), and the applicable appendices (Volume 3)

  8. Nuclear Dynamics Consequence Analysis (NDCA) for the Disposal of Spent Nuclear Fuel in an Underground Geologic Repository--Volume 2: Methodology and Results

    Energy Technology Data Exchange (ETDEWEB)

    Taylor, L.L.; Wilson, J.R.; Sanchez, L.C.; Aguilar, R.; Trellue, H.R.; Cochrane, K.; Rath, J.S.

    1998-10-01

    The US Department of Energy Office of Environmental Management's (DOE/EM's) National Spent Nuclear Fuel Program (NSNFP), through a collaboration between Sandia National Laboratories (SNL) and Idaho National Engineering and Environmental Laboratory (INEEL), is conducting a systematic Nuclear Dynamics Consequence Analysis (NDCA) of the disposal of SNFs in an underground geologic repository sited in unsaturated tuff. This analysis is intended to provide interim guidance to the DOE for the management of the SNF while they prepare for final compliance evaluation. This report presents results from a Nuclear Dynamics Consequence Analysis (NDCA) that examined the potential consequences and risks of criticality during the long-term disposal of spent nuclear fuel owned by DOE-EM. This analysis investigated the potential of post-closure criticality, the consequences of a criticality excursion, and the probability frequency for post-closure criticality. The results of the NDCA are intended to provide the DOE-EM with a technical basis for measuring risk which can be used for screening arguments to eliminate post-closure criticality FEPs (features, events and processes) from consideration in the compliance assessment because of either low probability or low consequences. This report is composed of an executive summary (Volume 1), the methodology and results of the NDCA (Volume 2), and the applicable appendices (Volume 3).

  9. Scanner calibration revisited

    Directory of Open Access Journals (Sweden)

    Pozhitkov Alexander E

    2010-07-01

    Full Text Available Abstract Background Calibration of a microarray scanner is critical for accurate interpretation of microarray results. Shi et al. (BMC Bioinformatics, 2005, 6, Art. No. S11 Suppl. 2. reported usage of a Full Moon BioSystems slide for calibration. Inspired by the Shi et al. work, we have calibrated microarray scanners in our previous research. We were puzzled however, that most of the signal intensities from a biological sample fell below the sensitivity threshold level determined by the calibration slide. This conundrum led us to re-investigate the quality of calibration provided by the Full Moon BioSystems slide as well as the accuracy of the analysis performed by Shi et al. Methods Signal intensities were recorded on three different microarray scanners at various photomultiplier gain levels using the same calibration slide from Full Moon BioSystems. Data analysis was conducted on raw signal intensities without normalization or transformation of any kind. Weighted least-squares method was used to fit the data. Results We found that initial analysis performed by Shi et al. did not take into account autofluorescence of the Full Moon BioSystems slide, which led to a grossly distorted microarray scanner response. Our analysis revealed that a power-law function, which is explicitly accounting for the slide autofluorescence, perfectly described a relationship between signal intensities and fluorophore quantities. Conclusions Microarray scanners respond in a much less distorted fashion than was reported by Shi et al. Full Moon BioSystems calibration slides are inadequate for performing calibration. We recommend against using these slides.

  10. The turning points of world history : financial and methodological interpretations

    OpenAIRE

    Kukliński, Antoni (ed.); Pawłowski, Krzysztof (ed.); Swianiewicz, Jan (ed.)

    2012-01-01

    Volumes VI and VII of the REUPUS Series “The Atlantic Community. The Titanic of the XXI Century?” and “The Turning Points of World History” can be seen as twin volumes trying to find new empirical observations, new methodological approaches and new value judgements to face the enigma of the XXI Century. Those volumes try to present some new interpretations of one of the greatest turning points of human history which is the essential feature of our times... Volume VII is the ...

  11. LLE Review Quarterly Report (April-June 2001). Volume 87

    Energy Technology Data Exchange (ETDEWEB)

    Hinterman, Thomas H. [Univ. of Rochester, NY (United States). Lab. for Laser Energetics

    2001-06-01

    This volume of the LLE Review, covering April–June 2001, features ''A Self-Calibrating, Multichannel Streak Camera for Inertial Confinement Fusion Applications'' by Dr. W. R. Donaldson, R. Boni, R. L. Keck, and P. A. Jaanimagi. This article (p. 109) describes the 60-beam streak camera system used on OMEGA and focuses on the hardware and software calibration techniques that maximize its utility. The system can diagnose each of the beams on every target shot and can measure beam energies with 8% accuracy and timing at 7 ps rms. Beam-to-beam power variations of less than 5% can be detected. Other articles in this volume are: Evolution of Shell Nonuniformities Near Peak Compression of a Spherical Implosion; Multibeam Stimulated Brillouin Scattering from Hot Solid-Target Plasmas; Hot-Electron Effect in Superconductors and Is Applications for Radiation Sensors; and, Scaling Law for Marginal Ignition.

  12. Commissioning and calibration of the ALICE-TPC

    Energy Technology Data Exchange (ETDEWEB)

    Wiechula, Jens

    2009-04-21

    ALICE (A Large Ion Collider Experiment), is the dedicated heavy-ion experiment at the Large Hadron Collider (LHC) at CERN. It is optimised to reconstruct and identify the particles created in a lead-lead collision with a centre of mass energy of 5.5 TeV. The main tracking detector is a large-volume time-projection chamber (TPC). With an active volume of about 88m{sup 3} and a total readout area of 32.5 m{sup 2} it is the most challenging TPC ever build. A central electrode divides the 5 m long detector into two drift regions. Each readout side is subdivided into 18 inner and 18 outer multi-wire proportional read-out chambers. The readout area is subdivided into 557568 pads, where each pad is read out by an electronic chain. A complex calibration is needed in order to reach the design position-resolution of the reconstructed particle tracks of about 200 {mu}m. The work at hand presents results of the pedestal and noise behaviour of the front-end electronics (FEE), measurements of the pulse-shaping properties of the FEE using results obtained with a calibration pulser and measurements performed with the laser-calibration system. The data concerned were taken during two phases of the TPC commissioning. First measurements were performed in the clean room where the TPC was built. After the TPC was moved underground and built into the experiment, a second round of commissioning took place. Noise measurements in the clean room revealed a very large fraction of pads with noise values larger than the design specifications. Two modifications helped to reduce the 'ground bounce' effect: A desynchronisation in the the start of the readout of groups of channels and a modification in the grounding scheme of the FEE. Further noise measurements were carried out after the TPC has been moved to the experimental area underground. Here even a larger fraction of channels showed too large noise values. To study the shaping properties of the FEE a calibration pulser was used

  13. Commissioning and calibration of the ALICE-TPC

    International Nuclear Information System (INIS)

    Wiechula, Jens

    2009-01-01

    ALICE (A Large Ion Collider Experiment), is the dedicated heavy-ion experiment at the Large Hadron Collider (LHC) at CERN. It is optimised to reconstruct and identify the particles created in a lead-lead collision with a centre of mass energy of 5.5 TeV. The main tracking detector is a large-volume time-projection chamber (TPC). With an active volume of about 88m 3 and a total readout area of 32.5 m 2 it is the most challenging TPC ever build. A central electrode divides the 5 m long detector into two drift regions. Each readout side is subdivided into 18 inner and 18 outer multi-wire proportional read-out chambers. The readout area is subdivided into 557568 pads, where each pad is read out by an electronic chain. A complex calibration is needed in order to reach the design position-resolution of the reconstructed particle tracks of about 200 μm. The work at hand presents results of the pedestal and noise behaviour of the front-end electronics (FEE), measurements of the pulse-shaping properties of the FEE using results obtained with a calibration pulser and measurements performed with the laser-calibration system. The data concerned were taken during two phases of the TPC commissioning. First measurements were performed in the clean room where the TPC was built. After the TPC was moved underground and built into the experiment, a second round of commissioning took place. Noise measurements in the clean room revealed a very large fraction of pads with noise values larger than the design specifications. Two modifications helped to reduce the 'ground bounce' effect: A desynchronisation in the the start of the readout of groups of channels and a modification in the grounding scheme of the FEE. Further noise measurements were carried out after the TPC has been moved to the experimental area underground. Here even a larger fraction of channels showed too large noise values. To study the shaping properties of the FEE a calibration pulser was used. To generate signals in

  14. An efficient multi-stage algorithm for full calibration of the hemodynamic model from BOLD signal responses

    KAUST Repository

    Zambri, Brian; Djellouli, Rabia; Laleg-Kirati, Taous-Meriem

    2017-01-01

    We propose a computational strategy that falls into the category of prediction/correction iterative-type approaches, for calibrating the hemodynamic model introduced by Friston et al. (2000). The proposed method is employed to estimate consecutively the values of the biophysiological system parameters and the external stimulus characteristics of the model. Numerical results corresponding to both synthetic and real functional Magnetic Resonance Imaging (fMRI) measurements for a single stimulus as well as for multiple stimuli are reported to highlight the capability of this computational methodology to fully calibrate the considered hemodynamic model. This article is protected by copyright. All rights reserved.

  15. An efficient multi-stage algorithm for full calibration of the hemodynamic model from BOLD signal responses

    KAUST Repository

    Zambri, Brian

    2017-02-22

    We propose a computational strategy that falls into the category of prediction/correction iterative-type approaches, for calibrating the hemodynamic model introduced by Friston et al. (2000). The proposed method is employed to estimate consecutively the values of the biophysiological system parameters and the external stimulus characteristics of the model. Numerical results corresponding to both synthetic and real functional Magnetic Resonance Imaging (fMRI) measurements for a single stimulus as well as for multiple stimuli are reported to highlight the capability of this computational methodology to fully calibrate the considered hemodynamic model. This article is protected by copyright. All rights reserved.

  16. Development of calibration facility for radon and its progenies at NIM (China)

    International Nuclear Information System (INIS)

    Liang, J.C.; Liu, H.R.; Zhang, M.; Zheng, P.H.; Guo, Q.J.; Yang, Z.J.; Li, Z.S.; Zhang, L.

    2015-01-01

    Accurate measurement of radon and its progenies is the basis to control the radon dose and reduce the risk of lung cancer caused. The precise calibration of measuring instrument is an important part of the quality control of measurements of the concentration of radon and radon progenies. To establish Chinese national standards and realise reliable calibrations of measuring instrument for radon and its progenies, a radon chamber with regulation capability of environmental parameters, aerosol and radon concentrations was designed and constructed at National Institute of Metrology (NIM). The chamber has a total volume of ∼20 m 3 including an exposure volume of 12.44 m 3 . The radon concentration can be controlled from 12 Bq m -3 to the maximum of 232 kBq m -3 . The regulation range of temperature, relative humidity and aerosol are 0.66-44.39 deg. C, 16.4-95 %RH and 10 2 -10 6 cm -3 , respectively. The main advantages of the NIM radon chamber with respect to maintaining a stable concentration and equilibrium factor of radon progenies in a wide range through automatic regulation and control of radon and aerosol are described. (authors)

  17. A Radon Chamber without Radium Source for Detector Calibration and Radon Measurements

    International Nuclear Information System (INIS)

    Al-Azmi, D.; Karunakara, N.

    2008-01-01

    A radon chamber of volume 216 liters was designed and constructed for calibration of radon detectors and radon test measurements. The main feature of this chamber is that the active 226 Ra source, to generate the 222 Rn inside the chamber volume, is not required. Instead, 222 Rn from soil gas is utilized for this purpose. The supply of radon comes from the soil gas. Soil gas is drawn from the soil to fill the chamber with high radon concentration levels (∼ 80 kBq/m3). Desired radon concentration levels can be obtained by drawing the soil gas for different time durations and/or flow rate (author)

  18. Absolute radiometric calibration of Landsat using a pseudo invariant calibration site

    Science.gov (United States)

    Helder, D.; Thome, K.J.; Mishra, N.; Chander, G.; Xiong, Xiaoxiong; Angal, A.; Choi, Tae-young

    2013-01-01

    Pseudo invariant calibration sites (PICS) have been used for on-orbit radiometric trending of optical satellite systems for more than 15 years. This approach to vicarious calibration has demonstrated a high degree of reliability and repeatability at the level of 1-3% depending on the site, spectral channel, and imaging geometries. A variety of sensors have used this approach for trending because it is broadly applicable and easy to implement. Models to describe the surface reflectance properties, as well as the intervening atmosphere have also been developed to improve the precision of the method. However, one limiting factor of using PICS is that an absolute calibration capability has not yet been fully developed. Because of this, PICS are primarily limited to providing only long term trending information for individual sensors or cross-calibration opportunities between two sensors. This paper builds an argument that PICS can be used more extensively for absolute calibration. To illustrate this, a simple empirical model is developed for the well-known Libya 4 PICS based on observations by Terra MODIS and EO-1 Hyperion. The model is validated by comparing model predicted top-of-atmosphere reflectance values to actual measurements made by the Landsat ETM+ sensor reflective bands. Following this, an outline is presented to develop a more comprehensive and accurate PICS absolute calibration model that can be Système international d'unités (SI) traceable. These initial concepts suggest that absolute calibration using PICS is possible on a broad scale and can lead to improved on-orbit calibration capabilities for optical satellite sensors.

  19. Nonlinear propagation model for ultrasound hydrophones calibration in the frequency range up to 100 MHz.

    Science.gov (United States)

    Radulescu, E G; Wójcik, J; Lewin, P A; Nowicki, A

    2003-06-01

    To facilitate the implementation and verification of the new ultrasound hydrophone calibration techniques described in the companion paper (somewhere in this issue) a nonlinear propagation model was developed. A brief outline of the theoretical considerations is presented and the model's advantages and disadvantages are discussed. The results of simulations yielding spatial and temporal acoustic pressure amplitude are also presented and compared with those obtained using KZK and Field II models. Excellent agreement between all models is evidenced. The applicability of the model in discrete wideband calibration of hydrophones is documented in the companion paper somewhere in this volume.

  20. Online neural trigger for optimizing data acquisition during particle beam calibration tests with calorimeters

    International Nuclear Information System (INIS)

    Silva, P.V.M. da; Seixas, J.M. de; Damazio, D.O.; Ferreira, B.C.

    2004-01-01

    For LHC, the hadronic calorimetry of the ATLAS detector is performed by Tilecal, a scintillating tile calorimeter. For calibration purposes, a fraction of the Tilecal modules is placed in particle beam lines. Despite beam high quality, experimental beam contamination is observed and this masks the actual performance of the calorimeter. For optimizing the calibration task, an online neural particle classifier was developed for Tilecal. Envisaging a neural trigger for incoming particles, a neural process runs integrated to the data acquisition task and performs online training for particle identification. The neural classification performance is evaluated by correlating the neural response to classical methodology, confirming an ability for outsider identification at levels as high as 99.3%

  1. Online neural trigger for optimizing data acquisition during particle beam calibration tests with calorimeters

    CERN Document Server

    Da Silva, P V M; De Seixas, J M; Ferreira, B C

    2004-01-01

    For LHC, the hadronic calorimetry of the ATLAS detector is performed by Tilecal, a scintillating tile calorimeter. For calibration purposes, a fraction of the Tilecal modules is placed in particle beam lines. Despite beam high quality, experimental beam contamination is observed and this masks the actual performance of the calorimeter. For optimizing the calibration task, an online neural particle classifier was developed for Tilecal. Envisaging a neural trigger for incoming particles, a neural process runs integrated to the data acquisition task and performs online training for particle identification. The neural classification performance is evaluated by correlating the neural response to classical methodology, confirming an ability for outsider identification at levels as high as 99.3%.

  2. Assessment of calibration parameters for an aerial gamma spectrometry system using Monte-Carlo technique

    International Nuclear Information System (INIS)

    Srinivasan, P.; Raman, Anand; Sharma, D.N.

    2009-01-01

    Aerial gamma spectrometry is a very effective method for quickly surveying a large area, which might get contaminated following a nuclear accident, or due to nuclear weapon fallout. The technique not only helps in identifying the contaminating radionuclide but also in assessing the magnitude and the extent of contamination. These two factors are of importance for the authorities to quickly plan and execute effective counter measures and controls if required. The development of Airborne gamma ray spectrometry systems have been reported by different institutions. The application of these systems have been reported by different authors. Radiation Safety Systems Division of the Bhabha Atomic Research Centre has developed an Aerial Gamma Spectrometry System (AGSS) and the surveying methodology. For an online assessment of the contamination levels, it is essential to calibrate the system (AGSS) either flying it over a known contaminated area or over a simulated contaminated surface by deploying sealed sources on the ground. AGSS has been calibrated for different detectors in aerial exercises using such simulated contamination on the ground. The calibration methodology essentially needs net photo-peak counts in selected energy windows to finally arrive at the Air to Ground Correlation Factors at selected flight parameters such as altitude, speed of flight and the time interval at which each spectrum is acquired. This paper describes the methodology to predict all the necessary parameters like photon fluence at various altitudes, the photo-peak counts in different energy windows, Air to Ground Correlation Factors(AGCF), the dose rate at any height due to air scattered gamma ray photons etc. These parameters are predicted for a given source deployment matrix, detector and altitude of flying using the Monte-Carlo code MCNP (Monte Carlo Neutron and Photon Transport Code.CCC-200, RSIC, ORNL, Tennessee, USA). A methodology to generate the completely folded gamma ray count

  3. Latent Trait Theory Applications to Test Item Bias Methodology. Research Memorandum No. 1.

    Science.gov (United States)

    Osterlind, Steven J.; Martois, John S.

    This study discusses latent trait theory applications to test item bias methodology. A real data set is used in describing the rationale and application of the Rasch probabilistic model item calibrations across various ethnic group populations. A high school graduation proficiency test covering reading comprehension, writing mechanics, and…

  4. Control volume based hydrocephalus research

    Science.gov (United States)

    Cohen, Benjamin; Voorhees, Abram; Wei, Timothy

    2008-11-01

    Hydrocephalus is a disease involving excess amounts of cerebral spinal fluid (CSF) in the brain. Recent research has shown correlations to pulsatility of blood flow through the brain. However, the problem to date has presented as too complex for much more than statistical analysis and understanding. This talk will highlight progress on developing a fundamental control volume approach to studying hydrocephalus. The specific goals are to select physiologically control volume(s), develop conservation equations along with the experimental capabilities to accurately quantify terms in those equations. To this end, an in vitro phantom is used as a simplified model of the human brain. The phantom's design consists of a rigid container filled with a compressible gel. The gel has a hollow spherical cavity representing a ventricle and a cylindrical passage representing the aquaducts. A computer controlled piston pump supplies pulsatile volume fluctuations into and out of the flow phantom. MRI is used to measure fluid velocity, and volume change as functions of time. Independent pressure measurements and flow rate measurements are used to calibrate the MRI data. These data are used as a framework for future work with live patients.

  5. Development of new methodology for dose calculation in photographic dosimetry

    International Nuclear Information System (INIS)

    Daltro, T.F.L.

    1994-01-01

    A new methodology for equivalent dose calculations has been developed at IPEN-CNEN/SP to be applied at the Photographic Dosimetry Laboratory using artificial intelligence techniques by means of neutral network. The research was orientated towards the optimization of the whole set of parameters involves in the film processing going from the irradiation in order to obtain the calibration curve up to the optical density readings. The learning of the neutral network was performed by taking the readings of optical density from calibration curve as input and the effective energy and equivalent dose as output. The obtained results in the intercomparison show an excellent agreement with the actual values of dose and energy given by the National Metrology Laboratory of Ionizing Radiation. (author)

  6. Development of new methodology for dose calculation in photographic dosimetry

    International Nuclear Information System (INIS)

    Daltro, T.F.L.; Campos, L.L.

    1994-01-01

    A new methodology for equivalent dose calculation has been developed at IPEN-CNEN/SP to be applied at the Photographic Dosimetry Laboratory using artificial intelligence techniques by means of neural network. The research was oriented towards the optimization of the whole set of parameters involved in the film processing going from the irradiation in order to obtain the calibration curve up to the optical density readings. The learning of the neural network was performed by taking readings of optical density from calibration curve as input and the effective energy and equivalent dose as output. The obtained results in the intercomparison show an excellent agreement with the actual values of dose and energy given by the National Metrology Laboratory of Ionizing Radiation

  7. Study and mitigation of calibration factor instabilities in a water vapor Raman lidar

    Directory of Open Access Journals (Sweden)

    L. David

    2017-07-01

    Full Text Available We have investigated calibration variations in the Rameau water vapor Raman lidar. This lidar system was developed by the Institut National de l'Information Géographique et Forestière (IGN together with the Laboratoire Atmosphères, Milieux, Observations Spatiales (LATMOS. It aims at calibrating Global Navigation Satellite System (GNSS measurements for tropospheric wet delays and sounding the water vapor variability in the lower troposphere. The Rameau system demonstrated good capacity in retrieving water vapor mixing ratio (WVMR profiles accurately in several campaigns. However, systematic short-term and long-term variations in the lidar calibration factor pointed to persistent instabilities. A careful testing of each subsystem independently revealed that these instabilities are mainly induced by mode fluctuations in the optic fiber used to couple the telescope to the detection subsystem and by the spatial nonuniformity of the photomultiplier photocathodes. Laboratory tests that replicate and quantify these instability sources are presented. A redesign of the detection subsystem is presented, which, combined with careful alignment procedures, is shown to significantly reduce the instabilities. Outdoor measurements were performed over a period of 5 months to check the stability of the modified lidar system. The calibration changes in the detection subsystem were monitored with lidar profile measurements using a common nitrogen filter in both Raman channels. A short-term stability of 2–3 % and a long-term drift of 2–3 % per month are demonstrated. Compared to the earlier Development of Methodologies for Water Vapour Measurement (DEMEVAP campaign, this is a 3-fold improvement in the long-term stability of the detection subsystem. The overall water vapor calibration factors were determined and monitored with capacitive humidity sensor measurements and with GPS zenith wet delay (ZWD data. The changes in the water vapor calibration factors

  8. Calibration transfer between electronic nose systems for rapid In situ measurement of pulp and paper industry emissions

    Energy Technology Data Exchange (ETDEWEB)

    Deshmukh, Sharvari [CSIR-National Environmental Engineering and Research Institute, Nagpur (India); Department of Instrumentation and Electronics Engineering, Jadavpur University, Kolkata (India); Kamde, Kalyani [CSIR-National Environmental Engineering and Research Institute, Nagpur (India); Jana, Arun [Center for Development of Advance Computing, Kolkata (India); Korde, Sanjivani [CSIR-National Environmental Engineering and Research Institute, Nagpur (India); Bandyopadhyay, Rajib [Department of Instrumentation and Electronics Engineering, Jadavpur University, Kolkata (India); Sankar, Ravi [Center for Development of Advance Computing, Kolkata (India); Bhattacharyya, Nabarun, E-mail: nabarun.bhattacharya@cdac.in [Center for Development of Advance Computing, Kolkata (India); Pandey, R.A., E-mail: ra_pandey@neeri.res.in [CSIR-National Environmental Engineering and Research Institute, Nagpur (India)

    2014-09-02

    Highlights: • E-nose developed for obnoxious emissions measurement at pulp and paper industrial site. • ANN model developed for prediction of (CH{sub 3}){sub 2}S, (CH{sub 3}){sub 2}S{sub 2}, CH{sub 3}SH and H{sub 2}S concentration. • Calibration transfer methodology developed for transfer between two e-nose instruments. • Box–Behnken design and robust regression used for calibration transfer. • Results show effective transfer of training model from one e-nose system to other. - Abstract: Electronic nose systems when deployed in network mesh can effectively provide a low budget and onsite solution for the industrial obnoxious gaseous measurement. For accurate and identical prediction capability by all the electronic nose systems, a reliable calibration transfer model needs to be implemented in order to overcome the inherent sensor array variability. In this work, robust regression (RR) is used for calibration transfer between two electronic nose systems using a Box–Behnken (BB) design. Out of the two electronic nose systems, one was trained using industrial gas samples by four artificial neural network models, for the measurement of obnoxious odours emitted from pulp and paper industries. The emissions constitute mainly of hydrogen sulphide (H{sub 2}S), methyl mercaptan (MM), dimethyl sulphide (DMS) and dimethyl disulphide (DMDS) in different proportions. A Box–Behnken design consisting of 27 experiment sets based on synthetic gas combinations of H{sub 2}S, MM, DMS and DMDS, were conducted for calibration transfer between two identical electronic nose systems. Identical sensors on both the systems were mapped and the prediction models developed using ANN were then transferred to the second system using BB–RR methodology. The results showed successful transmission of prediction models developed for one system to other system, with the mean absolute error between the actual and predicted concentration of analytes in mg L{sup −1} after calibration

  9. Calibration transfer between electronic nose systems for rapid In situ measurement of pulp and paper industry emissions

    International Nuclear Information System (INIS)

    Deshmukh, Sharvari; Kamde, Kalyani; Jana, Arun; Korde, Sanjivani; Bandyopadhyay, Rajib; Sankar, Ravi; Bhattacharyya, Nabarun; Pandey, R.A.

    2014-01-01

    Highlights: • E-nose developed for obnoxious emissions measurement at pulp and paper industrial site. • ANN model developed for prediction of (CH 3 ) 2 S, (CH 3 ) 2 S 2 , CH 3 SH and H 2 S concentration. • Calibration transfer methodology developed for transfer between two e-nose instruments. • Box–Behnken design and robust regression used for calibration transfer. • Results show effective transfer of training model from one e-nose system to other. - Abstract: Electronic nose systems when deployed in network mesh can effectively provide a low budget and onsite solution for the industrial obnoxious gaseous measurement. For accurate and identical prediction capability by all the electronic nose systems, a reliable calibration transfer model needs to be implemented in order to overcome the inherent sensor array variability. In this work, robust regression (RR) is used for calibration transfer between two electronic nose systems using a Box–Behnken (BB) design. Out of the two electronic nose systems, one was trained using industrial gas samples by four artificial neural network models, for the measurement of obnoxious odours emitted from pulp and paper industries. The emissions constitute mainly of hydrogen sulphide (H 2 S), methyl mercaptan (MM), dimethyl sulphide (DMS) and dimethyl disulphide (DMDS) in different proportions. A Box–Behnken design consisting of 27 experiment sets based on synthetic gas combinations of H 2 S, MM, DMS and DMDS, were conducted for calibration transfer between two identical electronic nose systems. Identical sensors on both the systems were mapped and the prediction models developed using ANN were then transferred to the second system using BB–RR methodology. The results showed successful transmission of prediction models developed for one system to other system, with the mean absolute error between the actual and predicted concentration of analytes in mg L −1 after calibration transfer (on second system) being 0.076, 0

  10. Assessing the Predictive Capability of the LIFEIV Nuclear Fuel Performance Code using Sequential Calibration

    International Nuclear Information System (INIS)

    Stull, Christopher J.; Williams, Brian J.; Unal, Cetin

    2012-01-01

    would be reduced considerably. The authors note that the PMI is primarily intended to provide a high-level, quantitative description of year-to-year (or version-to-version) improvements in numerical models, where these descriptions can be used as a means of justifying funding requests to support further model development research. It is in this context that the present report should be considered: the availability of data from experimental tests should be viewed as a time-dependent variable, where experiments are added to the calibration suite as resources become available. For the present report, the experimental data is of course already available (permitting demonstration of the proposed methodology). Furthermore, the authors are not proposing this methodology as the answer to the question of how to allocate resources for experimental tests, and readers are directed to (5) and the references contained in Section 1 of (5) for additional information on the subject. However, the strength of this methodology is that it offers a means by which to select the sequence of experiments in a pre-arranged experimental campaign (a situation for which the methods discussed in (5) are less appropriate). The report is organized as follows. Section 2 describes the methodology employed to formulate the sequences of experiments for the calibrations performed for this study. Section 3 then presents the results associated with two sequences; supplementary results are provided in the Appendix. The report then concludes in Section 4 with a brief summary.

  11. Methodology for calibration of detector of NaI (TI)) 3 ' X 3 ' for in vivo measurements of patients with hyperthyroidism undergoing to radioiodotherapy; Metodologia para calibracao de detector de NaI(TI) ) 3'X3' para medicoes in vivo em pacientes portadores de hipertireoidismo submetidos a radioiodoterapia

    Energy Technology Data Exchange (ETDEWEB)

    Carvalho, Carlaine B.; Lacerda, Isabelle V.B.; Oliveira, Mercia L.; Hazin, Clovis A., E-mail: carlaine.carvalho@gmail.com, E-mail: bellelacerda@hotmail.com, E-mail: mercial@cnen.gov.br, E-mail: chazin@cnen.gov.br [Universidade Federal de Pernambuco (DEN/UFPE), Recife, PE (Brazil). Departamento de Energia Nuclear; Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE/CNEN-PE), Recife, PE (Brazil); Lima, Fabiana F., E-mail: fflima@cnen.gov.br [Centro Regional de Ciencias Nucleares do Nordeste (CRCN-NE/CNEN-PE), Recife, PE (Brazil)

    2013-10-01

    The aim of this study is to establish the methodology for calibration of the detection system to be used in determining the therapeutic activity of {sup 131}I required to release desired absorbed dose in the thyroid gland . This step is critical to the development of a protocol for individualized doses. The system consists of a detector of NaI (Tl ) 3'x3' coupled to software Genie 2000. We used the calibration sources of {sup 60}Co , {sup 137}Cs and {sup 133}Ba. We obtained the straight calibration system, with sources {sup 60}Co and {sup 137}Cs. Subsequently , the detector was calibrated using a thyroid phantom-neck designed and produced by the IRD / CNEN with known activity of {sup 133}Ba standard solution containing 18.7 kBq (on 09/24/12) evenly distributed. He was also calibrated with other thyroid- neck phantom model 3108 manufactured by Searle Radigraphics Ind., containing a liquid source of {sup 131}I ( 7.7 MBq ). Five measurements were performed during 5 minutes for three different distances detector-simulator and calculated the corresponding calibration factors . The values of the calibration factors found for the simulator made by IRD and Searle Radigraphics Ind. for the distances 20, 25 and 30 cm were 0.35 , 0.24, 0.18, 0.15 , 0.11, 0, 09 , respectively. With the detection system properly calibrated and the calibration factors established, the technique is suitable for the evaluation of diagnostic activities of {sup 131}I incorporated by hyperthyroid patients. (author)

  12. Dataset for: An efficient multi-stage algorithm for full calibration of the hemodynamic model from BOLD signal responses

    KAUST Repository

    Djellouli, Rabia

    2017-01-01

    We propose a computational strategy that falls into the category of prediction/correction iterative-type approaches, for calibrating the hemodynamic model introduced by Friston et al. (2000). The proposed method is employed to estimate consecutively the values of the biophysiological system parameters and the external stimulus characteristics of the model. Numerical results corresponding to both synthetic and real functional Magnetic Resonance Imaging (fMRI) measurements for a single stimulus as well as for multiple stimuli are reported to highlight the capability of this computational methodology to fully calibrate the considered hemodynamic model.

  13. Calibration of radionuclides with decay trough beta emission or electron capture by liquid scintillation technique

    International Nuclear Information System (INIS)

    Loureiro, Jamir dos Santos

    2000-02-01

    In this work is reported a methodology a methodology for pure beta and electron capture radionuclides standardization, suing liquid scintillation technique. In this sense the CIEMAT/NIST method, recently utilized by international laboratories, was implemented and the lack in the Laboratorio Nacional das Radiacoes Ionizantes - LNMRI, of the Comissao Nacional de Energia Nuclear - CNEN, for adequate methodology to standardize this kind if radionuclides was filled, fact that was not present with alpha and gamma radionuclides. The implementation procedure evaluation was provided by concentration activity determination of the following radionuclides: 14 C and 90 Sr, pure beta emitters; 55 Fe, electron capture decay; 204 Tl, electron capture and beta decay and 60 Co, beta-gamma emitter. In this way, a careful analysis of the implementation procedure with these radionuclides types, ranging on a broad energy spectral, was possible. To check the calibration results, intercomparisons among our measurements of these radionuclides and the reference values of the CIEMAT/Spain laboratory were provided. To check the calibration results, intercomparisons among our measurements of these radionuclides and the reference values of the CIEMAT/Spain laboratory were provided. Besides this intercomparisons, one was provided with a 204 Tl solution, utilized in the international comparison recently promoted by BIPM, and another one with a 60 C solution calibrated in LNMRI/CNEN previously by a relative calibration system, with a well type pressurized ionization chamber, and an absolute beta-gamma coincidence system, with a pill-box type proportional counter 4 π geometry, coupled with a scintillator system with a sodium iodide cristal of 4x4 inches. The comparisons among LNMRI/CNEN results and the reference values, showed a small deviation of 1,32% for 14 C, 0,40% for 60 Co, 1,12% for 55 Fe, 0,10% for 90 Sr and 0,73% for 204 Tl. For the BIPM solution the deviation was 0,46% and for 60 Co

  14. Calibration-on-the-spot”: How to calibrate an EMCCD camera from its images

    DEFF Research Database (Denmark)

    Mortensen, Kim; Flyvbjerg, Henrik

    2016-01-01

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction-......-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information....

  15. ISS Payload Racks Automated Flow Control Calibration Method

    Science.gov (United States)

    Simmonds, Boris G.

    2003-01-01

    Payload Racks utilize MTL and/or LTL station water for cooling of payloads and avionics. Flow control range from valves of fully closed, to up to 300 Ibmhr. Instrument accuracies are as high as f 7.5 Ibm/hr for flow sensors and f 3 Ibm/hr for valve controller, for a total system accuracy of f 10.5 Ibm/hr. Improved methodology was developed, tested and proven that reduces accuracy of the commanded flows to less than f 1 Ibmhr. Uethodology could be packed in a "calibration kit" for on- orbit flow sensor checkout and recalibration, extending the rack operations before return to earth. -

  16. Calibration of the hadronic calorimeter prototype for a future lepton collider

    Energy Technology Data Exchange (ETDEWEB)

    Schroeder, Sarah; Garutti, Erika [Institute for Experimental Physics, Hamburg University, Luruper Chaussee 149, D-22761 Hamburg (Germany); Collaboration: CALICE-D-Collaboration

    2016-07-01

    The CALICE AHCAL technological prototype is a hadronic calorimeter prototype for a future e{sup +}e{sup -} - collider. It is designed as a sampling calorimeter alternating steel absorber plates and active readout layers, segmented in single plastic scintillator tiles of 3 x 3 x 0.3 cm{sup 3} volume. Each tile is individually coupled to a silicon photomultiplier, read out by a dedicated ASIC with energy measurement and time stamping capability. The high granularity is meant to enable imaging and separation of single showers, for a Particle Flow approach to the jet energy measurement. The prototype aims to establish a scalable solution for an ILC detector. A total of 3456 calorimeter cells need to be inter-calibrated, for this the response to muons is used. The calibration procedure is presented, and the statistic and systematic uncertainties are discussed, which have a direct impact on the constant term of the calorimeter energy resolution. Additionally, the MIP yield in number of fired SiPM pixels can be compared betw een the muon calibration and a test bench calibrations obtained using a Sr sourc e on the single tiles before the assembly of the calorimeter. A good correlation would enable pre-calibation of the single channels on the test bench to be port able to the assemble detector. This hypothesis is checked with the present work.

  17. Calibration of reference KAP-meters at SSDL and cross calibration of clinical KAP-meters

    International Nuclear Information System (INIS)

    Hetland, Per O.; Friberg, Eva G.; Oevreboe, Kirsti M.; Bjerke, Hans H.

    2009-01-01

    In the summer of 2007 the secondary standard dosimetry laboratory (SSDL) in Norway established a calibration service for reference air-kerma product meter (KAP-meter). The air-kerma area product, PKA, is a dosimetric quantity that can be directly related to the patient dose and used for risk assessment associated with different x-ray examinations. The calibration of reference KAP-meters at the SSDL gives important information on parameters influencing the calibration factor for different types of KAP-meters. The use of reference KAP-meters calibrated at the SSDL is an easy and reliable way to calibrate or verify the PKA indicated by the x-ray equipment out in the clinics. Material and methods. Twelve KAP-meters were calibrated at the SSDL by use of the substitution method at five diagnostic radiation qualities (RQRs). Results. The calibration factors varied from 0.94 to 1.18. The energy response of the individual KAP-meters varied by a total of 20% between the different RQRs and the typical chamber transmission factors ranged from 0.78 to 0.91. Discussion. It is important to use a calibrated reference KAP-meter and a harmonised calibration method in the PKA calibration in hospitals. The obtained uncertainty in the PKA readings is comparable with other calibration methods if the information in the calibration certificate is correct used, corrections are made and proper positioning of the KAP-chamber is performed. This will ensure a reliable estimate of the patient dose and a proper optimisation of conventional x-ray examinations and interventional procedures

  18. Atlas based brain volumetry: How to distinguish regional volume changes due to biological or physiological effects from inherent noise of the methodology.

    Science.gov (United States)

    Opfer, Roland; Suppa, Per; Kepp, Timo; Spies, Lothar; Schippling, Sven; Huppertz, Hans-Jürgen

    2016-05-01

    Fully-automated regional brain volumetry based on structural magnetic resonance imaging (MRI) plays an important role in quantitative neuroimaging. In clinical trials as well as in clinical routine multiple MRIs of individual patients at different time points need to be assessed longitudinally. Measures of inter- and intrascanner variability are crucial to understand the intrinsic variability of the method and to distinguish volume changes due to biological or physiological effects from inherent noise of the methodology. To measure regional brain volumes an atlas based volumetry (ABV) approach was deployed using a highly elastic registration framework and an anatomical atlas in a well-defined template space. We assessed inter- and intrascanner variability of the method in 51 cognitively normal subjects and 27 Alzheimer dementia (AD) patients from the Alzheimer's Disease Neuroimaging Initiative by studying volumetric results of repeated scans for 17 compartments and brain regions. Median percentage volume differences of scan-rescans from the same scanner ranged from 0.24% (whole brain parenchyma in healthy subjects) to 1.73% (occipital lobe white matter in AD), with generally higher differences in AD patients as compared to normal subjects (e.g., 1.01% vs. 0.78% for the hippocampus). Minimum percentage volume differences detectable with an error probability of 5% were in the one-digit percentage range for almost all structures investigated, with most of them being below 5%. Intrascanner variability was independent of magnetic field strength. The median interscanner variability was up to ten times higher than the intrascanner variability. Copyright © 2016 Elsevier Inc. All rights reserved.

  19. Reachable volume RRT

    KAUST Repository

    McMahon, Troy

    2015-05-01

    © 2015 IEEE. Reachable volumes are a new technique that allows one to efficiently restrict sampling to feasible/reachable regions of the planning space even for high degree of freedom and highly constrained problems. However, they have so far only been applied to graph-based sampling-based planners. In this paper we develop the methodology to apply reachable volumes to tree-based planners such as Rapidly-Exploring Random Trees (RRTs). In particular, we propose a reachable volume RRT called RVRRT that can solve high degree of freedom problems and problems with constraints. To do so, we develop a reachable volume stepping function, a reachable volume expand function, and a distance metric based on these operations. We also present a reachable volume local planner to ensure that local paths satisfy constraints for methods such as PRMs. We show experimentally that RVRRTs can solve constrained problems with as many as 64 degrees of freedom and unconstrained problems with as many as 134 degrees of freedom. RVRRTs can solve problems more efficiently than existing methods, requiring fewer nodes and collision detection calls. We also show that it is capable of solving difficult problems that existing methods cannot.

  20. Reachable volume RRT

    KAUST Repository

    McMahon, Troy; Thomas, Shawna; Amato, Nancy M.

    2015-01-01

    © 2015 IEEE. Reachable volumes are a new technique that allows one to efficiently restrict sampling to feasible/reachable regions of the planning space even for high degree of freedom and highly constrained problems. However, they have so far only been applied to graph-based sampling-based planners. In this paper we develop the methodology to apply reachable volumes to tree-based planners such as Rapidly-Exploring Random Trees (RRTs). In particular, we propose a reachable volume RRT called RVRRT that can solve high degree of freedom problems and problems with constraints. To do so, we develop a reachable volume stepping function, a reachable volume expand function, and a distance metric based on these operations. We also present a reachable volume local planner to ensure that local paths satisfy constraints for methods such as PRMs. We show experimentally that RVRRTs can solve constrained problems with as many as 64 degrees of freedom and unconstrained problems with as many as 134 degrees of freedom. RVRRTs can solve problems more efficiently than existing methods, requiring fewer nodes and collision detection calls. We also show that it is capable of solving difficult problems that existing methods cannot.

  1. Methodologies and intelligent systems for technology enhanced learning

    CERN Document Server

    Gennari, Rosella; Vitorini, Pierpaolo; Vicari, Rosa; Prieta, Fernando

    2014-01-01

    This volume presents recent research on Methodologies and Intelligent Systems for Technology Enhanced Learning. It contains the contributions of ebuTEL 2013 conference which took place in Trento, Italy, on September, 16th 2013 and of mis4TEL 2014 conference, which took take place in Salamanca, Spain, on September, 4th-6th 2014 This conference series are an open forum for discussing intelligent systems for Technology Enhanced Learning and empirical methodologies for its design or evaluation.

  2. Automated calibration of TECAN genesis liquid handling workstation utilizing an online balance and density meter.

    Science.gov (United States)

    Xie, Iris H; Wang, Michael H; Carpenter, Richard; Wu, Henry Y

    2004-02-01

    With robotics widely used in bioanalytical assays, accurate system performance is essential to ensure the quality and productivity of the robotics. In our lab, an automated calibration procedure has been developed to evaluate the precision and accuracy of the TECAN (Research Triangle Park, NC, U.S.A.) Genesis liquid handling system in a bioanalytical laboratory setting. The calibrations were performed by transferring and weighing the solvents automatically on a microbalance controlled by a Gemini program. From the data acquired, calibration reports were generated using a template. The novel aspect of this approach is the use of an on-line balance and a density meter, both of which combine to make the calibration process simple, efficient, and precise. For quantitative bioanalysis, a variety of solvents, including methanol, water, mixed solvents, and plasma, are typically used to prepare standards and unknown samples. Density information is usually unknown for the mixed solvents, and the density of plasma can vary from species to species. However, with the use of a universal density meter, the density could be obtained in seconds. The issue of solvent evaporation during the calibration process was also addressed. Calibration curves were set up for various liquid classes. Pipetting volumes ranged from 10 microL to 900 microL. Precision and accuracy results obtained from the semiannual performance evaluations showed this procedure to be reliable and user-friendly. Using the automated calibration procedure, the calibration and performance evaluation of the robotic system is considerably more efficient, and the incidence of unacceptable precision and accuracy is greatly reduced.

  3. An internally consistent pressure calibration of geobarometers applicable to the Earth’s upper mantle using in situ XRD

    Energy Technology Data Exchange (ETDEWEB)

    Beyer, Christopher; Rosenthal, Anja; Myhill, Robert; Crichton, Wilson A.; Yu, Tony; Wang, Yanbin; Frost, Daniel J.

    2018-02-01

    We have performed an experimental cross calibration of a suite of mineral equilibria within mantle rock bulk compositions that are commonly used in geobarometry to determine the equilibration depths of upper mantle assemblages. Multiple barometers were compared simultaneously in experimental runs, where the pressure was determined using in-situ measurements of the unit cell volumes of MgO, NaCl, Re and h-BN between 3.6 and 10.4 GPa, and 1250 and 1500 °C. The experiments were performed in a large volume press (LVPs) in combination with synchrotron X-ray diffraction. Noble metal capsules drilled with multiple sample chambers were loaded with a range of bulk compositions representative of peridotite, eclogite and pyroxenite lithologies. By this approach, we simultaneously calibrated the geobarometers applicable to different mantle lithologies under identical and well determined pressure and temperature conditions. We identified discrepancies between the calculated and experimental pressures for which we propose simple linear or constant correction factors to some of the previously published barometric equations. As a result, we establish internally-consistent cross-calibrations for a number of garnet-orthopyroxene, garnet-clinopyroxene, Ca-Tschermaks-in-clinopyroxene and majorite geobarometers.

  4. Calibration of regional palaeohydrogeology and sensitivity analysis using hydrochemistry data in site investigations

    International Nuclear Information System (INIS)

    Hunter, F.M.I.; Hartley, L.J.; Hoch, A.; Jackson, C.P.; McCarthy, R.; Marsic, N.; Gylling, B.

    2008-01-01

    A transient coupled regional model of groundwater flow and solute transport has been developed, which allows the use of hydrochemical data to calibrate the model input parameters. The methodology has been illustrated using examples from the Simpevarp area in south-eastern Sweden which is being considered for geological disposal of spent nuclear fuel. The 3-dimensional model includes descriptions of spatial heterogeneity, density driven flow, rock matrix diffusion and transport and mixing of different water types, and has been simulated between 8000 BC and 2000 AD. Present-day analyses of major elemental ions and stable isotopes have been used to calibrate the model, which has then been cross checked against measured hydraulic conductivities, and against the hydrochemical interpretation of reference water mixing fractions. The key hydrogeological model sensitivities have been identified using the calibrated model and are found to include high sensitivity to the top surface flow boundary condition, the influence of variations in fracture transmissivity in different orientations (anisotropy), spatial heterogeneity in the deterministic regional deformation zones and the spacing between water bearing fractures (in terms of its effect on matrix diffusion)

  5. Assessment of annual pollutant loads in combined sewers from continuous turbidity measurements: sensitivity to calibration data.

    Science.gov (United States)

    Lacour, C; Joannis, C; Chebbo, G

    2009-05-01

    This article presents a methodology for assessing annual wet weather Suspended Solids (SS) and Chemical Oxygen Demand (COD) loads in combined sewers, along with the associated uncertainties from continuous turbidity measurements. The proposed method is applied to data from various urban catchments in the cities of Paris and Nantes. The focus here concerns the impact of the number of rain events sampled for calibration (i.e. through establishing linear SS/turbidity or COD/turbidity relationships) on the uncertainty of annual pollutant load assessments. Two calculation methods are investigated, both of which rely on Monte Carlo simulations: random assignment of event-specific calibration relationships to each individual rain event, and the use of an overall relationship built from the entire available data set. Since results indicate a fairly low inter-event variability for calibration relationship parameters, an accurate assessment of pollutant loads can be derived, even when fewer than 10 events are sampled for calibration purposes. For operational applications, these results suggest that turbidity could provide a more precise evaluation of pollutant loads at lower cost than typical sampling methods.

  6. Investigation on calibration parameter of mammography calibration facilities at MINT

    International Nuclear Information System (INIS)

    Asmaliza Hashim; Wan Hazlinda Ismail; Md Saion Salikin; Muhammad Jamal Md Isa; Azuhar Ripin; Norriza Mohd Isa

    2004-01-01

    A mammography calibration facility has been established in the Medical Physics Laboratory, Malaysian Institute for Nuclear Technology Research (MINT). The calibration facility is established at the national level mainly to provide calibration services for radiation measuring test instruments or test tools used in quality assurance programme in mammography, which is being implemented in Malaysia. One of the accepted parameters that determine the quality of a radiation beam is the homogeneity coefficient. It is determined from the values of the 1 st and 2 nd Half Value Layer (HVL). In this paper, the consistency of the mammography machine beam qualities that is available in MINT, is investigated and presented. For calibration purposes, five radiation qualities namely 23, 25, 28, 30 and 35 kV, selectable from the control panel of the X-ray machine is used. Important parameters that are set for this calibration facility are exposure time, tube current, focal spot to detector distance (FDD) and beam size at specific distance. The values of homogeneity coefficient of this laboratory for the past few years tip to now be presented in this paper. Backscatter radiations are also considered in this investigation. (Author)

  7. Calibration and statistical techniques for building an interactive screen for learning of alphabets by children

    Directory of Open Access Journals (Sweden)

    Riby Abraham Boby

    2017-05-01

    Full Text Available This article focuses on the implementation details of a portable interactive device called Image-projective Desktop Varnamala Trainer. The device uses a projector to produce a virtual display on a flat surface. For enabling interaction, the information about a user’s hand movement is obtained from a single two-dimensional scanning laser range finder in contrast with a camera sensor used in many earlier applications. A generalized calibration process to obtain exact transformation from projected screen coordinate system to sensor coordinate system is proposed in this article and implemented for enabling interaction. This permits production of large interactive displays with minimal cost. Additionally, it makes the entire system portable, that is, display can be produced on any planar surface like floor, tabletop, and so on. The calibration and its performance have been evaluated by varying screen sizes and the number of points used for calibration. The device was successfully calibrated for different screens. A novel learning-based methodology for predicting a user’s behaviour was then realized to improve the system’s performance. This has been experimentally evaluated, and the overall accuracy of prediction was about 96%. An application was then designed for this set-up to improve the learning of alphabets by the children through an interactive audiovisual feedback system. It uses a game-based methodology to help students learn in a fun way. Currently, it has bilingual (Hindi and English user interface to enable learning of alphabets and elementary mathematics. A user survey was conducted after demonstrating it to school children. The survey results are very encouraging. Additionally, a study to ascertain the improvement in the learning outcome of the children was done. The results clearly indicate an improvement in the learning outcome of the children who used the device over those who did not.

  8. A scattering methodology for droplet sizing of e-cigarette aerosols.

    Science.gov (United States)

    Pratte, Pascal; Cosandey, Stéphane; Goujon-Ginglinger, Catherine

    2016-10-01

    Knowledge of the droplet size distribution of inhalable aerosols is important to predict aerosol deposition yield at various respiratory tract locations in human. Optical methodologies are usually preferred over the multi-stage cascade impactor for high-throughput measurements of aerosol particle/droplet size distributions. Evaluate the Laser Aerosol Spectrometer technology based on Polystyrene Sphere Latex (PSL) calibration curve applied for the experimental determination of droplet size distributions in the diameter range typical of commercial e-cigarette aerosols (147-1361 nm). This calibration procedure was tested for a TSI Laser Aerosol Spectrometer (LAS) operating at a wavelength of 633 nm and assessed against model di-ethyl-hexyl-sebacat (DEHS) droplets and e-cigarette aerosols. The PSL size response was measured, and intra- and between-day standard deviations calculated. DEHS droplet sizes were underestimated by 15-20% by the LAS when the PSL calibration curve was used; however, the intra- and between-day relative standard deviations were e-cigarette aerosols ranged from 130-191 nm to 225-293 nm, respectively, similar to published values. The LAS instrument can be used to measure e-cigarette aerosol droplet size distributions with a bias underestimating the expected value by 15-20% when using a precise PSL calibration curve. Controlled variability of DEHS size measurements can be achieved with the LAS system; however, this method can only be applied to test aerosols having a refractive index close to that of PSL particles used for calibration.

  9. Improved best estimate plus uncertainty methodology including advanced validation concepts to license evolving nuclear reactors

    International Nuclear Information System (INIS)

    Unal, Cetin; Williams, Brian; McClure, Patrick; Nelson, Ralph A.

    2010-01-01

    Many evolving nuclear energy programs plan to use advanced predictive multi-scale multi-physics simulation and modeling capabilities to reduce cost and time from design through licensing. Historically, the role of experiments was primary tool for design and understanding of nuclear system behavior while modeling and simulation played the subordinate role of supporting experiments. In the new era of multi-scale multi-physics computational based technology development, the experiments will still be needed but they will be performed at different scales to calibrate and validate models leading predictive simulations. Cost saving goals of programs will require us to minimize the required number of validation experiments. Utilization of more multi-scale multi-physics models introduces complexities in the validation of predictive tools. Traditional methodologies will have to be modified to address these arising issues. This paper lays out the basic aspects of a methodology that can be potentially used to address these new challenges in design and licensing of evolving nuclear technology programs. The main components of the proposed methodology are verification, validation, calibration, and uncertainty quantification. An enhanced calibration concept is introduced and is accomplished through data assimilation. The goal is to enable best-estimate prediction of system behaviors in both normal and safety related environments. To achieve this goal requires the additional steps of estimating the domain of validation and quantification of uncertainties that allow for extension of results to areas of the validation domain that are not directly tested with experiments, which might include extension of the modeling and simulation (M and S) capabilities for application to full-scale systems. The new methodology suggests a formalism to quantify an adequate level of validation (predictive maturity) with respect to required selective data so that required testing can be minimized for

  10. Improved best estimate plus uncertainty methodology including advanced validation concepts to license evolving nuclear reactors

    Energy Technology Data Exchange (ETDEWEB)

    Unal, Cetin [Los Alamos National Laboratory; Williams, Brian [Los Alamos National Laboratory; Mc Clure, Patrick [Los Alamos National Laboratory; Nelson, Ralph A [IDAHO NATIONAL LAB

    2010-01-01

    Many evolving nuclear energy programs plan to use advanced predictive multi-scale multi-physics simulation and modeling capabilities to reduce cost and time from design through licensing. Historically, the role of experiments was primary tool for design and understanding of nuclear system behavior while modeling and simulation played the subordinate role of supporting experiments. In the new era of multi-scale multi-physics computational based technology development, the experiments will still be needed but they will be performed at different scales to calibrate and validate models leading predictive simulations. Cost saving goals of programs will require us to minimize the required number of validation experiments. Utilization of more multi-scale multi-physics models introduces complexities in the validation of predictive tools. Traditional methodologies will have to be modified to address these arising issues. This paper lays out the basic aspects of a methodology that can be potentially used to address these new challenges in design and licensing of evolving nuclear technology programs. The main components of the proposed methodology are verification, validation, calibration, and uncertainty quantification. An enhanced calibration concept is introduced and is accomplished through data assimilation. The goal is to enable best-estimate prediction of system behaviors in both normal and safety related environments. To achieve this goal requires the additional steps of estimating the domain of validation and quantification of uncertainties that allow for extension of results to areas of the validation domain that are not directly tested with experiments, which might include extension of the modeling and simulation (M&S) capabilities for application to full-scale systems. The new methodology suggests a formalism to quantify an adequate level of validation (predictive maturity) with respect to required selective data so that required testing can be minimized for cost

  11. Real options and volume uncertainty by field development projects

    International Nuclear Information System (INIS)

    Ekern, S.; Stensland, G.

    1993-12-01

    The report concerns a study on the use of option methodology in field development projects. The report shows how the value of flexibility in the different decision processes is to be found by means of real option methodology. Particular attention is laid on the uncertainty concerning the volume of reserves and production capacity. The results from the study were based on the research project dubbed ''Use of real options in field development projects''. The project is partially connected to another project dubbed ''Decisive behaviour and alternative action under uncertainty in the petroleum sector''. Main topics cover as follow: Example with volume uncertainty; real options and volume uncertainty; gradual disclosure of uncertainty in the production; value of flexible production equipment. 33 refs., 19 figs., 17 tabs

  12. Air kerma standard for calibration of well-type chambers in Brazil using {sup 192}Ir HDR sources and its traceability

    Energy Technology Data Exchange (ETDEWEB)

    Di Prinzio, Renato; Almeida, Carlos Eduardo de [Laboratorio de Ciencias Radiologicas-Universidade do Estado do Rio de Janeiro (LCR/UERJ), R. Sao Francisco Xavier, 524, Pavilhao Haroldo Lisboa da Cunha, Terreo, Sala 136-Maracana, CEP 20550-900-Rio de Janeiro/RJ-Rio de Janeiro, RJ (Brazil) and Instituto de Radioprotecao e Dosimetria-Comissao Nacional de Energia Nuclear (IRD/CNEN), Av. Salvador Allende, s/n, Jacarepagua-CE22780-160-Rio de Janeiro, RJ (Brazil); Laboratorio de Ciencias Radiologicas-Universidade do Estado do Rio de Janeiro (LCR/UERJ), R. Sao Francisco Xavier, 524, Pavilhao Haroldo Lisboa da Cunha, Terreo, Sala 136-Maracana, CEP 20550-900-Rio de Janeiro/RJ-Rio de Janeiro, RJ (Brazil)

    2009-03-15

    In Brazil there are over 100 high dose rate (HDR) brachytherapy facilities using well-type chambers for the determination of the air kerma rate of {sup 192}Ir sources. This paper presents the methodology developed and extensively tested by the Laboratorio de Ciencias Radiologicas (LCR) and presently in use to calibrate those types of chambers. The system was initially used to calibrate six well-type chambers of brachytherapy services, and the maximum deviation of only 1.0% was observed between the calibration coefficients obtained and the ones in the calibration certificate provided by the UWADCL. In addition to its traceability to the Brazilian National Standards, the whole system was taken to University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL) for a direct comparison and the same formalism to calculate the air kerma was used. The comparison results between the two laboratories show an agreement of 0.9% for the calibration coefficients. Three Brazilian well-type chambers were calibrated at the UWADCL, and by LCR, in Brazil, using the developed system and a clinical HDR machine. The results of the calibration of three well chambers have shown an agreement better than 1.0%. Uncertainty analyses involving the measurements made both at the UWADCL and LCR laboratories are discussed.

  13. Wavelength calibration of imaging spectrometer using atmospheric absorption features

    Science.gov (United States)

    Zhou, Jiankang; Chen, Yuheng; Chen, Xinhua; Ji, Yiqun; Shen, Weimin

    2012-11-01

    Imaging spectrometer is a promising remote sensing instrument widely used in many filed, such as hazard forecasting, environmental monitoring and so on. The reliability of the spectral data is the determination to the scientific communities. The wavelength position at the focal plane of the imaging spectrometer will change as the pressure and temperature vary, or the mechanical vibration. It is difficult for the onboard calibration instrument itself to keep the spectrum reference accuracy and it also occupies weight and the volume of the remote sensing platform. Because the spectral images suffer from the atmospheric effects, the carbon oxide, water vapor, oxygen and solar Fraunhofer line, the onboard wavelength calibration can be processed by the spectral images themselves. In this paper, wavelength calibration is based on the modeled and measured atmospheric absorption spectra. The modeled spectra constructed by the atmospheric radiative transfer code. The spectral angle is used to determine the best spectral similarity between the modeled spectra and measured spectra and estimates the wavelength position. The smile shape can be obtained when the matching process across all columns of the data. The present method is successful applied on the Hyperion data. The value of the wavelength shift is obtained by shape matching of oxygen absorption feature and the characteristics are comparable to that of the prelaunch measurements.

  14. Calibration Under Uncertainty.

    Energy Technology Data Exchange (ETDEWEB)

    Swiler, Laura Painton; Trucano, Timothy Guy

    2005-03-01

    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  15. AN AROMATIC INVENTORY OF THE LOCAL VOLUME

    International Nuclear Information System (INIS)

    Marble, A. R.; Engelbracht, C. W.; Block, M.; Van Zee, L.; Dale, D. A.; Cohen, S. A.; Schuster, M. D.; Smith, J. D. T.; Gordon, K. D.; Wu, Y.; Lee, J. C.; Kennicutt, R. C.; Skillman, E. D.; Johnson, L. C.; Calzetti, D.; Lee, H.

    2010-01-01

    Using infrared photometry from the Spitzer Space Telescope, we perform the first inventory of aromatic feature emission (also commonly referred to as polycyclic aromatic hydrocarbon emission) for a statistically complete sample of star-forming galaxies in the local volume. The photometric methodology involved is calibrated and demonstrated to recover the aromatic fraction of the Infrared Array Camera 8 μm flux with a standard deviation of 6% for a training set of 40 SINGS galaxies (ranging from stellar to dust dominated) with both suitable mid-infrared Spitzer Infrared Spectrograph spectra and equivalent photometry. A potential factor of 2 improvement could be realized with suitable 5.5 μm and 10 μm photometry, such as what may be provided in the future by the James Webb Space Telescope. The resulting technique is then applied to mid-infrared photometry for the 258 galaxies from the Local Volume Legacy (LVL) survey, a large sample dominated in number by low-luminosity dwarf galaxies for which obtaining comparable mid-infrared spectroscopy is not feasible. We find the total LVL luminosity due to five strong aromatic features in the 8 μm complex to be 2.47 x 10 10 L sun with a mean volume density of 8.8 x 10 6 L sun Mpc -3 . Twenty-four of the LVL galaxies, corresponding to a luminosity cut at M B = -18.22, account for 90% of the aromatic luminosity. Using oxygen abundances compiled from the literature for 129 of the 258 LVL galaxies, we find a correlation between metallicity and the aromatic-to-total infrared emission ratio but not the aromatic-to-total 8 μm dust emission ratio. A possible explanation is that metallicity plays a role in the abundance of aromatic molecules relative to the total dust content, but other factors, such as star formation and/or the local radiation field, affect the excitation of those molecules.

  16. Calibration of reconstruction parameters in atom probe tomography using a single crystallographic orientation

    International Nuclear Information System (INIS)

    Suram, Santosh K.; Rajan, Krishna

    2013-01-01

    The purpose of this work is to develop a methodology to estimate the APT reconstruction parameters when limited crystallographic information is available. Reliable spatial scaling of APT data currently requires identification of multiple crystallographic poles from the field desorption image for estimating the reconstruction parameters. This requirement limits the capacity of accurately reconstructing APT data for certain complex systems, such as highly alloyed systems and nanostructured materials wherein more than one pole is usually not observed within one grain. To overcome this limitation, we develop a quantitative methodology for calibrating the reconstruction parameters in an APT dataset by ensuring accurate inter-planar spacing and optimizing the curvature correction for the atomic planes corresponding to a single crystallographic orientation. We validate our approach on an aluminum dataset and further illustrate its capabilities by computing geometric reconstruction parameters for W and Al–Mg–Sc datasets. - Highlights: ► Quantitative approach is developed to accurately reconstruct APT data. ► Curvature of atomic planes in APT data is used to calibrate the reconstruction. ► APT reconstruction parameters are determined from a single crystallographic axis. ► Quantitative approach is demonstrated on W, Al and Al–Mg–Sc systems. ► Accurate APT reconstruction of complex materials is now possible

  17. Researching virtual worlds methodologies for studying emergent practices

    CERN Document Server

    Phillips, Louise

    2013-01-01

    This volume presents a wide range of methodological strategies that are designed to take into account the complex, emergent, and continually shifting character of virtual worlds. It interrogates how virtual worlds emerge as objects of study through the development and application of various methodological strategies. Virtual worlds are not considered objects that exist as entities with fixed attributes independent of our continuous engagement with them and interpretation of them. Instead, they are conceived of as complex ensembles of technology, humans, symbols, discourses, and economic structures, ensembles that emerge in ongoing practices and specific situations. A broad spectrum of perspectives and methodologies is presented: Actor-Network-Theory and post-Actor-Network-Theory, performativity theory, ethnography, discourse analysis, Sense-Making Methodology, visual ethnography, multi-sited ethnography, and Social Network Analysis.

  18. Calibration of Flick standards

    International Nuclear Information System (INIS)

    Thalmann, Ruedi; Spiller, Jürg; Küng, Alain; Jusko, Otto

    2012-01-01

    Flick standards or magnification standards are widely used for an efficient and functional calibration of the sensitivity of form measuring instruments. The results of a recent measurement comparison have shown to be partially unsatisfactory and revealed problems related to the calibration of these standards. In this paper the influence factors for the calibration of Flick standards using roundness measurement instruments are discussed in detail, in particular the bandwidth of the measurement chain, residual form errors of the device under test, profile distortions due to the diameter of the probing element and questions related to the definition of the measurand. The different contributions are estimated using simulations and are experimentally verified. Also alternative methods to calibrate Flick standards are investigated. Finally the practical limitations of Flick standard calibration are shown and the usability of Flick standards both to calibrate the sensitivity of roundness instruments and to check the filter function of such instruments is analysed. (paper)

  19. Tolerance limits and tolerance intervals for ratios of normal random variables using a bootstrap calibration.

    Science.gov (United States)

    Flouri, Marilena; Zhai, Shuyan; Mathew, Thomas; Bebu, Ionut

    2017-05-01

    This paper addresses the problem of deriving one-sided tolerance limits and two-sided tolerance intervals for a ratio of two random variables that follow a bivariate normal distribution, or a lognormal/normal distribution. The methodology that is developed uses nonparametric tolerance limits based on a parametric bootstrap sample, coupled with a bootstrap calibration in order to improve accuracy. The methodology is also adopted for computing confidence limits for the median of the ratio random variable. Numerical results are reported to demonstrate the accuracy of the proposed approach. The methodology is illustrated using examples where ratio random variables are of interest: an example on the radioactivity count in reverse transcriptase assays and an example from the area of cost-effectiveness analysis in health economics. © 2017 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  20. Benefit-Cost Analysis of Integrated Paratransit Systems : Volume 6. Technical Appendices.

    Science.gov (United States)

    1979-09-01

    This last volume, includes five technical appendices which document the methodologies used in the benefit-cost analysis. They are the following: Scenario analysis methodology; Impact estimation; Example of impact estimation; Sensitivity analysis; Agg...

  1. Synthesis Polarimetry Calibration

    Science.gov (United States)

    Moellenbrock, George

    2017-10-01

    Synthesis instrumental polarization calibration fundamentals for both linear (ALMA) and circular (EVLA) feed bases are reviewed, with special attention to the calibration heuristics supported in CASA. Practical problems affecting modern instruments are also discussed.

  2. CALIBRATION OF LOW COST DIGITAL CAMERA USING DATA FROM SIMULTANEOUS LIDAR AND PHOTOGRAMMETRIC SURVEYS

    Directory of Open Access Journals (Sweden)

    E. Mitishita

    2012-07-01

    Full Text Available Digital photogrammetric products from the integration of imagery and lidar datasets are a reality nowadays. When the imagery and lidar surveys are performed together and the camera is connected to the lidar system, a direct georeferencing can be applied to compute the exterior orientation parameters of the images. Direct georeferencing of the images requires accurate interior orientation parameters to perform photogrammetric application. Camera calibration is a procedure applied to compute the interior orientation parameters (IOPs. Calibration researches have established that to obtain accurate IOPs, the calibration must be performed with same or equal condition that the photogrammetric survey is done. This paper shows the methodology and experiments results from in situ self-calibration using a simultaneous images block and lidar dataset. The calibration results are analyzed and discussed. To perform this research a test field was fixed in an urban area. A set of signalized points was implanted on the test field to use as the check points or control points. The photogrammetric images and lidar dataset of the test field were taken simultaneously. Four strips of flight were used to obtain a cross layout. The strips were taken with opposite directions of flight (W-E, E-W, N-S and S-N. The Kodak DSC Pro SLR/c digital camera was connected to the lidar system. The coordinates of the exposition station were computed from the lidar trajectory. Different layouts of vertical control points were used in the calibration experiments. The experiments use vertical coordinates from precise differential GPS survey or computed by an interpolation procedure using the lidar dataset. The positions of the exposition stations are used as control points in the calibration procedure to eliminate the linear dependency of the group of interior and exterior orientation parameters. This linear dependency happens, in the calibration procedure, when the vertical images and

  3. Sandia WIPP calibration traceability

    Energy Technology Data Exchange (ETDEWEB)

    Schuhen, M.D. [Sandia National Labs., Albuquerque, NM (United States); Dean, T.A. [RE/SPEC, Inc., Albuquerque, NM (United States)

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities.

  4. Sandia WIPP calibration traceability

    International Nuclear Information System (INIS)

    Schuhen, M.D.; Dean, T.A.

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities

  5. Calibration of a user-defined mine blast model in LSDYNA and comparison with ale simultions

    NARCIS (Netherlands)

    Verreault, J.; Leerdam, P.J.C.; Weerheijm, J.

    2016-01-01

    The calibration of a user-defined blast model implemented in LS-DYNA is presented using full-scale test rig experiments, partly according to the NATO STANAG 4569 AEP-55 Volume 2 specifications where the charge weight varies between 6 kg and 10 kg and the burial depth is 100 mm and deeper. The model

  6. Virtual Distances Methodology as Verification Technique for AACMMs with a Capacitive Sensor Based Indexed Metrology Platform

    Science.gov (United States)

    Acero, Raquel; Santolaria, Jorge; Brau, Agustin; Pueo, Marcos

    2016-01-01

    This paper presents a new verification procedure for articulated arm coordinate measuring machines (AACMMs) together with a capacitive sensor-based indexed metrology platform (IMP) based on the generation of virtual reference distances. The novelty of this procedure lays on the possibility of creating virtual points, virtual gauges and virtual distances through the indexed metrology platform’s mathematical model taking as a reference the measurements of a ball bar gauge located in a fixed position of the instrument’s working volume. The measurements are carried out with the AACMM assembled on the IMP from the six rotating positions of the platform. In this way, an unlimited number and types of reference distances could be created without the need of using a physical gauge, therefore optimizing the testing time, the number of gauge positions and the space needed in the calibration and verification procedures. Four evaluation methods are presented to assess the volumetric performance of the AACMM. The results obtained proved the suitability of the virtual distances methodology as an alternative procedure for verification of AACMMs using the indexed metrology platform. PMID:27869722

  7. Virtual Distances Methodology as Verification Technique for AACMMs with a Capacitive Sensor Based Indexed Metrology Platform.

    Science.gov (United States)

    Acero, Raquel; Santolaria, Jorge; Brau, Agustin; Pueo, Marcos

    2016-11-18

    This paper presents a new verification procedure for articulated arm coordinate measuring machines (AACMMs) together with a capacitive sensor-based indexed metrology platform (IMP) based on the generation of virtual reference distances. The novelty of this procedure lays on the possibility of creating virtual points, virtual gauges and virtual distances through the indexed metrology platform's mathematical model taking as a reference the measurements of a ball bar gauge located in a fixed position of the instrument's working volume. The measurements are carried out with the AACMM assembled on the IMP from the six rotating positions of the platform. In this way, an unlimited number and types of reference distances could be created without the need of using a physical gauge, therefore optimizing the testing time, the number of gauge positions and the space needed in the calibration and verification procedures. Four evaluation methods are presented to assess the volumetric performance of the AACMM. The results obtained proved the suitability of the virtual distances methodology as an alternative procedure for verification of AACMMs using the indexed metrology platform.

  8. POLCAL - POLARIMETRIC RADAR CALIBRATION

    Science.gov (United States)

    Vanzyl, J.

    1994-01-01

    Calibration of polarimetric radar systems is a field of research in which great progress has been made over the last few years. POLCAL (Polarimetric Radar Calibration) is a software tool intended to assist in the calibration of Synthetic Aperture Radar (SAR) systems. In particular, POLCAL calibrates Stokes matrix format data produced as the standard product by the NASA/Jet Propulsion Laboratory (JPL) airborne imaging synthetic aperture radar (AIRSAR). POLCAL was designed to be used in conjunction with data collected by the NASA/JPL AIRSAR system. AIRSAR is a multifrequency (6 cm, 24 cm, and 68 cm wavelength), fully polarimetric SAR system which produces 12 x 12 km imagery at 10 m resolution. AIRSTAR was designed as a testbed for NASA's Spaceborne Imaging Radar program. While the images produced after 1991 are thought to be calibrated (phase calibrated, cross-talk removed, channel imbalance removed, and absolutely calibrated), POLCAL can and should still be used to check the accuracy of the calibration and to correct it if necessary. Version 4.0 of POLCAL is an upgrade of POLCAL version 2.0 released to AIRSAR investigators in June, 1990. New options in version 4.0 include automatic absolute calibration of 89/90 data, distributed target analysis, calibration of nearby scenes with calibration parameters from a scene with corner reflectors, altitude or roll angle corrections, and calibration of errors introduced by known topography. Many sources of error can lead to false conclusions about the nature of scatterers on the surface. Errors in the phase relationship between polarization channels result in incorrect synthesis of polarization states. Cross-talk, caused by imperfections in the radar antenna itself, can also lead to error. POLCAL reduces cross-talk and corrects phase calibration without the use of ground calibration equipment. Removing the antenna patterns during SAR processing also forms a very important part of the calibration of SAR data. Errors in the

  9. A general methodology for three-dimensional analysis of variation in target volume delineation

    NARCIS (Netherlands)

    Remeijer, P.; Rasch, C.; Lebesque, J. V.; van Herk, M.

    1999-01-01

    A generic method for three-dimensional (3-D) evaluation of target volume delineation in multiple imaging modalities is presented. The evaluation includes geometrical and statistical methods to estimate observer differences and variability in defining the Gross Tumor Volume (GTV) in relation to the

  10. Canonical duality theory unified methodology for multidisciplinary study

    CERN Document Server

    Latorre, Vittorio; Ruan, Ning

    2017-01-01

    This book on canonical duality theory provides a comprehensive review of its philosophical origin, physics foundation, and mathematical statements in both finite- and infinite-dimensional spaces. A ground-breaking methodological theory, canonical duality theory can be used for modeling complex systems within a unified framework and for solving a large class of challenging problems in multidisciplinary fields in engineering, mathematics, and the sciences. This volume places a particular emphasis on canonical duality theory’s role in bridging the gap between non-convex analysis/mechanics and global optimization.  With 18 total chapters written by experts in their fields, this volume provides a nonconventional theory for unified understanding of the fundamental difficulties in large deformation mechanics, bifurcation/chaos in nonlinear science, and the NP-hard problems in global optimization. Additionally, readers will find a unified methodology and powerful algorithms for solving challenging problems in comp...

  11. Proposal of a postal system for Ir-192 sources calibration used in high dose rate brachytherapy with LiF:Mn:Ti thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Vieira, W.S.; Borges, J.C.; Almeida, C.E.V.

    1998-01-01

    A proposal in order to improve the brachytherapy quality control and to allow postal intercomparison of Ir-192 sources used in high dose rate brachytherapy has been presented. The LiF: Mn: Ti (TLD 100) detector has been selected for such purpose. The experimental array and the TLDs irradiation and calibration techniques, at the treatment units, have been specified in the light of more recent methodology of Ir-192 calibration sources. (Author)

  12. Geometric Calibration and Radiometric Correction of LiDAR Data and Their Impact on the Quality of Derived Products

    Directory of Open Access Journals (Sweden)

    Wai-Yeung Yan

    2011-09-01

    Full Text Available LiDAR (Light Detection And Ranging systems are capable of providing 3D positional and spectral information (in the utilized spectrum range of the mapped surface. Due to systematic errors in the system parameters and measurements, LiDAR systems require geometric calibration and radiometric correction of the intensity data in order to maximize the benefit from the collected positional and spectral information. This paper presents a practical approach for the geometric calibration of LiDAR systems and radiometric correction of collected intensity data while investigating their impact on the quality of the derived products. The proposed approach includes the use of a quasi-rigorous geometric calibration and the radar equation for the radiometric correction of intensity data. The proposed quasi-rigorous calibration procedure requires time-tagged point cloud and trajectory position data, which are available to most of the data users. The paper presents a methodology for evaluating the impact of the geometric calibration on the relative and absolute accuracy of the LiDAR point cloud. Furthermore, the impact of the geometric calibration and radiometric correction on land cover classification accuracy is investigated. The feasibility of the proposed methods and their impact on the derived products are demonstrated through experimental results using real data.

  13. Mathematical calibration procedure of a capacitive sensor-based indexed metrology platform

    International Nuclear Information System (INIS)

    Brau-Avila, A; Valenzuela-Galvan, M; Herrera-Jimenez, V M; Santolaria, J; Aguilar, J J; Acero, R

    2017-01-01

    The demand for faster and more reliable measuring tasks for the control and quality assurance of modern production systems has created new challenges for the field of coordinate metrology. Thus, the search for new solutions in coordinate metrology systems and the need for the development of existing ones still persists. One example of such a system is the portable coordinate measuring machine (PCMM), the use of which in industry has considerably increased in recent years, mostly due to its flexibility for accomplishing in-line measuring tasks as well as its reduced cost and operational advantages compared to traditional coordinate measuring machines. Nevertheless, PCMMs have a significant drawback derived from the techniques applied in the verification and optimization procedures of their kinematic parameters. These techniques are based on the capture of data with the measuring instrument from a calibrated gauge object, fixed successively in various positions so that most of the instrument measuring volume is covered, which results in time-consuming, tedious and expensive verification and optimization procedures. In this work the mathematical calibration procedure of a capacitive sensor-based indexed metrology platform (IMP) is presented. This calibration procedure is based on the readings and geometric features of six capacitive sensors and their targets with nanometer resolution. The final goal of the IMP calibration procedure is to optimize the geometric features of the capacitive sensors and their targets in order to use the optimized data in the verification procedures of PCMMs. (paper)

  14. Mathematical calibration procedure of a capacitive sensor-based indexed metrology platform

    Science.gov (United States)

    Brau-Avila, A.; Santolaria, J.; Acero, R.; Valenzuela-Galvan, M.; Herrera-Jimenez, V. M.; Aguilar, J. J.

    2017-03-01

    The demand for faster and more reliable measuring tasks for the control and quality assurance of modern production systems has created new challenges for the field of coordinate metrology. Thus, the search for new solutions in coordinate metrology systems and the need for the development of existing ones still persists. One example of such a system is the portable coordinate measuring machine (PCMM), the use of which in industry has considerably increased in recent years, mostly due to its flexibility for accomplishing in-line measuring tasks as well as its reduced cost and operational advantages compared to traditional coordinate measuring machines. Nevertheless, PCMMs have a significant drawback derived from the techniques applied in the verification and optimization procedures of their kinematic parameters. These techniques are based on the capture of data with the measuring instrument from a calibrated gauge object, fixed successively in various positions so that most of the instrument measuring volume is covered, which results in time-consuming, tedious and expensive verification and optimization procedures. In this work the mathematical calibration procedure of a capacitive sensor-based indexed metrology platform (IMP) is presented. This calibration procedure is based on the readings and geometric features of six capacitive sensors and their targets with nanometer resolution. The final goal of the IMP calibration procedure is to optimize the geometric features of the capacitive sensors and their targets in order to use the optimized data in the verification procedures of PCMMs.

  15. Water calibration measurements for neutron radiography: Application to water content quantification in porous media

    Energy Technology Data Exchange (ETDEWEB)

    Kang, M., E-mail: kangm@ornl.gov [Department of Earth and Planetary Sciences, University of Tennessee, Knoxville, TN (United States); Chemical and Engineering Materials Division, Oak Ridge National Laboratory, Oak Ridge, TN (United States); Bilheux, H.Z., E-mail: bilheuxhn@ornl.gov [Chemical and Engineering Materials Division, Oak Ridge National Laboratory, Oak Ridge, TN (United States); Voisin, S. [Computational Sciences and Engineering Division, Oak Ridge National Laboratory, Oak Ridge, TN (United States); Cheng, C.L.; Perfect, E. [Department of Earth and Planetary Sciences, University of Tennessee, Knoxville, TN (United States); Horita, J. [Department of Geosciences, Texas Tech University, Lubbock, TX (United States); Warren, J.M. [Environmental Sciences Division, Oak Ridge National Laboratory, Oak Ridge, TN (United States)

    2013-04-21

    Using neutron radiography, the measurement of water thickness was performed using aluminum (Al) water calibration cells at the High Flux Isotope Reactor (HFIR) Cold-Guide (CG) 1D neutron imaging facility at Oak Ridge National Laboratory, Oak Ridge, TN, USA. Calibration of water thickness is an important step to accurately measure water contents in samples of interest. Neutron attenuation by water does not vary linearly with thickness mainly due to beam hardening and scattering effects. Transmission measurements for known water thicknesses in water calibration cells allow proper correction of the underestimation of water content due to these effects. As anticipated, strong scattering effects were observed for water thicknesses greater than 0.2 cm when the water calibration cells were positioned close to the face of the detector/scintillator (0 and 2.4 cm away, respectively). The water calibration cells were also positioned 24 cm away from the detector face. These measurements resulted in less scattering and this position (designated as the sample position) was used for the subsequent experimental determination of the neutron attenuation coefficient for water. Neutron radiographic images of moist Flint sand in rectangular and cylindrical containers acquired at the sample position were used to demonstrate the applicability of the water calibration. Cumulative changes in the water volumes within the sand columns during monotonic drainage determined by neutron radiography were compared with those recorded by direct reading from a burette connected to a hanging water column. In general, the neutron radiography data showed very good agreement with those obtained volumetrically using the hanging water-column method. These results allow extension of the calibration equation to the quantification of unknown water contents within other samples of porous media.

  16. Calibration of hydrodynamic behavior and biokinetics for TOC removal modeling in biofilm reactors under different hydraulic conditions.

    Science.gov (United States)

    Zeng, Ming; Soric, Audrey; Roche, Nicolas

    2013-09-01

    In this study, total organic carbon (TOC) biodegradation was simulated by GPS-X software in biofilm reactors with carriers of plastic rings and glass beads under different hydraulic conditions. Hydrodynamic model by retention time distribution and biokinetic measurement by in-situ batch test served as two significant parts of model calibration. Experimental results showed that TOC removal efficiency was stable in both media due to the enough height of column, although the actual hydraulic volume changed during the variation of hydraulic condition. Simulated TOC removal efficiencies were close to experimental ones with low theil inequality coefficient values (below 0.15). Compared with glass beads, more TOC was removed in the filter with plastic rings due to the larger actual hydraulic volume and lower half saturation coefficient in spite of its lower maximum specific growth rate of biofilm, which highlighted the importance of calibrating hydrodynamic behavior and biokinetics. Copyright © 2013 Elsevier Ltd. All rights reserved.

  17. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  18. Calibrated Properties Model

    International Nuclear Information System (INIS)

    Ahlers, C.; Liu, H.

    2000-01-01

    The purpose of this Analysis/Model Report (AMR) is to document the Calibrated Properties Model that provides calibrated parameter sets for unsaturated zone (UZ) flow and transport process models for the Yucca Mountain Site Characterization Project (YMP). This work was performed in accordance with the ''AMR Development Plan for U0035 Calibrated Properties Model REV00. These calibrated property sets include matrix and fracture parameters for the UZ Flow and Transport Model (UZ Model), drift seepage models, drift-scale and mountain-scale coupled-processes models, and Total System Performance Assessment (TSPA) models as well as Performance Assessment (PA) and other participating national laboratories and government agencies. These process models provide the necessary framework to test conceptual hypotheses of flow and transport at different scales and predict flow and transport behavior under a variety of climatic and thermal-loading conditions

  19. Informal uncertainty analysis (GLUE of continuous flow simulation in a hybrid sewer system with infiltration inflow – consistency of containment ratios in calibration and validation?

    Directory of Open Access Journals (Sweden)

    A. Breinholt

    2013-10-01

    Full Text Available Monitoring of flows in sewer systems is increasingly applied to calibrate urban drainage models used for long-term simulation. However, most often models are calibrated without considering the uncertainties. The generalized likelihood uncertainty estimation (GLUE methodology is here applied to assess parameter and flow simulation uncertainty using a simplified lumped sewer model that accounts for three separate flow contributions: wastewater, fast runoff from paved areas, and slow infiltrating water from permeable areas. Recently GLUE methodology has been critisised for generating prediction limits without statistical coherence and consistency and for the subjectivity in the choice of a threshold value to distinguish "behavioural" from "non-behavioural" parameter sets. In this paper we examine how well the GLUE methodology performs when the behavioural parameter sets deduced from a calibration period are applied to generate prediction bounds in validation periods. By retaining an increasing number of parameter sets we aim at obtaining consistency between the GLUE generated 90% prediction limits and the actual containment ratio (CR in calibration. Due to the large uncertainties related to spatio-temporal rain variability during heavy convective rain events, flow measurement errors, possible model deficiencies as well as epistemic uncertainties, it was not possible to obtain an overall CR of more than 80%. However, the GLUE generated prediction limits still proved rather consistent, since the overall CRs obtained in calibration corresponded well with the overall CRs obtained in validation periods for all proportions of retained parameter sets evaluated. When focusing on wet and dry weather periods separately, some inconsistencies were however found between calibration and validation and we address here some of the reasons why we should not expect the coverage of the prediction limits to be identical in calibration and validation periods in real

  20. Detector characterization for efficiency calibration in different measurement geometries

    International Nuclear Information System (INIS)

    Toma, M.; Dinescu, L.; Sima, O.

    2005-01-01

    In order to perform an accurate efficiency calibration for different measurement geometries a good knowledge of the detector characteristics is required. The Monte Carlo simulation program GESPECOR is applied. The detector characterization required for Monte Carlo simulation is achieved using the efficiency values obtained from measuring a point source. The point source was measured in two significant geometries: the source placed in a vertical plane containing the vertical symmetry axis of the detector and in a horizontal plane containing the centre of the active volume of the detector. The measurements were made using gamma spectrometry technique. (authors)

  1. VOLUMNECT: measuring volumes with Kinect

    Science.gov (United States)

    Quintino Ferreira, Beatriz; Griné, Miguel; Gameiro, Duarte; Costeira, João. Paulo; Sousa Santos, Beatriz

    2014-03-01

    This article presents a solution to volume measurement object packing using 3D cameras (such as the Microsoft KinectTM). We target application scenarios, such as warehouses or distribution and logistics companies, where it is important to promptly compute package volumes, yet high accuracy is not pivotal. Our application auto- matically detects cuboid objects using the depth camera data and computes their volume and sorting it allowing space optimization. The proposed methodology applies to a point cloud simple computer vision and image processing methods, as connected components, morphological operations and Harris corner detector, producing encouraging results, namely an accuracy in volume measurement of 8mm. Aspects that can be further improved are identified; nevertheless, the current solution is already promising turning out to be cost effective for the envisaged scenarios.

  2. Application of virtual distances methodology to laser tracker verification with an indexed metrology platform

    International Nuclear Information System (INIS)

    Acero, R; Pueo, M; Santolaria, J; Aguilar, J J; Brau, A

    2015-01-01

    High-range measuring equipment like laser trackers need large dimension calibrated reference artifacts in their calibration and verification procedures. In this paper, a new verification procedure for portable coordinate measuring instruments based on the generation and evaluation of virtual distances with an indexed metrology platform is developed. This methodology enables the definition of an unlimited number of reference distances without materializing them in a physical gauge to be used as a reference. The generation of the virtual points and reference lengths derived is linked to the concept of the indexed metrology platform and the knowledge of the relative position and orientation of its upper and lower platforms with high accuracy. It is the measuring instrument together with the indexed metrology platform one that remains still, rotating the virtual mesh around them. As a first step, the virtual distances technique is applied to a laser tracker in this work. The experimental verification procedure of the laser tracker with virtual distances is simulated and further compared with the conventional verification procedure of the laser tracker with the indexed metrology platform. The results obtained in terms of volumetric performance of the laser tracker proved the suitability of the virtual distances methodology in calibration and verification procedures for portable coordinate measuring instruments, broadening and expanding the possibilities for the definition of reference distances in these procedures. (paper)

  3. Application of virtual distances methodology to laser tracker verification with an indexed metrology platform

    Science.gov (United States)

    Acero, R.; Santolaria, J.; Pueo, M.; Aguilar, J. J.; Brau, A.

    2015-11-01

    High-range measuring equipment like laser trackers need large dimension calibrated reference artifacts in their calibration and verification procedures. In this paper, a new verification procedure for portable coordinate measuring instruments based on the generation and evaluation of virtual distances with an indexed metrology platform is developed. This methodology enables the definition of an unlimited number of reference distances without materializing them in a physical gauge to be used as a reference. The generation of the virtual points and reference lengths derived is linked to the concept of the indexed metrology platform and the knowledge of the relative position and orientation of its upper and lower platforms with high accuracy. It is the measuring instrument together with the indexed metrology platform one that remains still, rotating the virtual mesh around them. As a first step, the virtual distances technique is applied to a laser tracker in this work. The experimental verification procedure of the laser tracker with virtual distances is simulated and further compared with the conventional verification procedure of the laser tracker with the indexed metrology platform. The results obtained in terms of volumetric performance of the laser tracker proved the suitability of the virtual distances methodology in calibration and verification procedures for portable coordinate measuring instruments, broadening and expanding the possibilities for the definition of reference distances in these procedures.

  4. White Sands Missile Range 2007 Urban Study: Data Processing - Volume DP-1 (Sonic Calibration)

    Science.gov (United States)

    2008-09-01

    calibrated in a wind tunnel against a NIST standard were positioned at the two ends of the lineup , closest to the buildings (positions 1 and 10 in...utilized westerly winds between 8 and 10 m/s (2007 February 24, 0600–0605 LT). The 5 min of data utilized by the “fast” case were sequential . The...2.4–11.2 m/s. The 5-min “variable” case utilized sequential minutes. Results from all three WS conditions were the following (Vaucher, 2007a

  5. PFMCal : Photonic force microscopy calibration extended for its application in high-frequency microrheology

    Science.gov (United States)

    Butykai, A.; Domínguez-García, P.; Mor, F. M.; Gaál, R.; Forró, L.; Jeney, S.

    2017-11-01

    The present document is an update of the previously published MatLab code for the calibration of optical tweezers in the high-resolution detection of the Brownian motion of non-spherical probes [1]. In this instance, an alternative version of the original code, based on the same physical theory [2], but focused on the automation of the calibration of measurements using spherical probes, is outlined. The new added code is useful for high-frequency microrheology studies, where the probe radius is known but the viscosity of the surrounding fluid maybe not. This extended calibration methodology is automatic, without the need of a user's interface. A code for calibration by means of thermal noise analysis [3] is also included; this is a method that can be applied when using viscoelastic fluids if the trap stiffness is previously estimated [4]. The new code can be executed in MatLab and using GNU Octave. Program Files doi:http://dx.doi.org/10.17632/s59f3gz729.1 Licensing provisions: GPLv3 Programming language: MatLab 2016a (MathWorks Inc.) and GNU Octave 4.0 Operating system: Linux and Windows. Supplementary material: A new document README.pdf includes basic running instructions for the new code. Journal reference of previous version: Computer Physics Communications, 196 (2015) 599 Does the new version supersede the previous version?: No. It adds alternative but compatible code while providing similar calibration factors. Nature of problem (approx. 50-250 words): The original code uses a MatLab-provided user's interface, which is not available in GNU Octave, and cannot be used outside of a proprietary software as MatLab. Besides, the process of calibration when using spherical probes needs an automatic method when calibrating big amounts of different data focused to microrheology. Solution method (approx. 50-250 words): The new code can be executed in the latest version of MatLab and using GNU Octave, a free and open-source alternative to MatLab. This code generates an

  6. Technical quality assessment of an optoelectronic system for movement analysis

    International Nuclear Information System (INIS)

    Sapienza University of Rome (Italy))" data-affiliation=" (Department of Mechanical and Aerospace Engineering, Sapienza University of Rome (Italy))" >Di Marco, R; Sapienza University of Rome (Italy))" data-affiliation=" (Department of Mechanical and Aerospace Engineering, Sapienza University of Rome (Italy))" >Patanè, F; Sapienza University of Rome (Italy))" data-affiliation=" (Department of Mechanical and Aerospace Engineering, Sapienza University of Rome (Italy))" >Cappa, P; Rossi, S

    2015-01-01

    The Optoelectronic Systems (OS) are largely used in gait analysis to evaluate the motor performances of healthy subjects and patients. The accuracy of marker trajectories reconstruction depends on several aspects: the number of cameras, the dimension and position of the calibration volume, and the chosen calibration procedure. In this paper we propose a methodology to evaluate the effects of the mentioned sources of error on the reconstruction of marker trajectories. The novel contribution of the present work consists in the dimension of the tested calibration volumes, which is comparable with the ones normally used in gait analysis; in addition, to simulate trajectories during clinical gait analysis, we provide non-default paths for markers as inputs. Several calibration procedures are implemented and the same trial is processed with each calibration file, also considering different cameras configurations. The RMSEs between the measured trajectories and the optimal ones are calculated for each comparison. To investigate the significant differences between the computed indices, an ANOVA analysis is implemented. The RMSE is sensible to the variations of the considered calibration volume and the camera configurations and it is always inferior to 43 mm

  7. METHODOLOGY OF THE DRUGS MARKET VOLUME MODELING ON THE EXAMPLE OF HEMOPHILIA A

    Directory of Open Access Journals (Sweden)

    N. B. Molchanova

    2015-01-01

    Full Text Available Hemophilia A is a serious genetic disease, which may lead to disability of a patient even in early ages without a required therapy. The only one therapeutic approach is a replacement therapy with drugs of bloodcoagulation factor VIII (FVIII. The modeling of coagulation drugs market volume will allow evaluation of the level of patients’ provision with a necessary therapy. Modeling of a “perfect” market of drugs and its comparison with the real one was the purpose of the study. During the modeling of market volume we have used the data about the number of hamophilia A patients on the basis of the federal registry, Russian and international morbidity indices, and the data of a real practice about average consumption of drugs of bloodcoagulation factors and data about the drugs prescription according to the standards and protocols of assistance rendering. According to the standards of care delivery, average annual volume of FVIII drugs consumption amounted to 406 325 244 IU for children and 964 578 678 IU for adults, i.e. an average volume of a “perfect” market is equal to 1 370 903 922 IU for all patients. The market volume is 1.8 times bigger than a real volume of FVIII drugs which, according to the data of IMS marketing agency, amounted to 765 000 000 IU in 2013. The modeling conducted has shown that despite a relatively high patients’ coverage there is a potential for almost double growth.

  8. The analytical calibration in (bio)imaging/mapping of the metallic elements in biological samples--definitions, nomenclature and strategies: state of the art.

    Science.gov (United States)

    Jurowski, Kamil; Buszewski, Bogusław; Piekoszewski, Wojciech

    2015-01-01

    Nowadays, studies related to the distribution of metallic elements in biological samples are one of the most important issues. There are many articles dedicated to specific analytical atomic spectrometry techniques used for mapping/(bio)imaging the metallic elements in various kinds of biological samples. However, in such literature, there is a lack of articles dedicated to reviewing calibration strategies, and their problems, nomenclature, definitions, ways and methods used to obtain quantitative distribution maps. The aim of this article was to characterize the analytical calibration in the (bio)imaging/mapping of the metallic elements in biological samples including (1) nomenclature; (2) definitions, and (3) selected and sophisticated, examples of calibration strategies with analytical calibration procedures applied in the different analytical methods currently used to study an element's distribution in biological samples/materials such as LA ICP-MS, SIMS, EDS, XRF and others. The main emphasis was placed on the procedures and methodology of the analytical calibration strategy. Additionally, the aim of this work is to systematize the nomenclature for the calibration terms: analytical calibration, analytical calibration method, analytical calibration procedure and analytical calibration strategy. The authors also want to popularize the division of calibration methods that are different than those hitherto used. This article is the first work in literature that refers to and emphasizes many different and complex aspects of analytical calibration problems in studies related to (bio)imaging/mapping metallic elements in different kinds of biological samples. Copyright © 2014 Elsevier B.V. All rights reserved.

  9. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR

    International Nuclear Information System (INIS)

    Acosta, Andy L. Romero; Lores, Stefan Gutierrez

    2013-01-01

    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba

  10. Primary calibration in acoustics metrology

    International Nuclear Information System (INIS)

    Milhomem, T A Bacelar; Soares, Z M Defilippo

    2015-01-01

    SI unit in acoustics is realized by the reciprocity calibrations of laboratory standard microphones in pressure field, free field and diffuse field. Calibrations in pressure field and in free field are already consolidated and the Inmetro already done them. Calibration in diffuse field is not yet consolidated, however, some national metrology institutes, including Inmetro, are conducting researches on this subject. This paper presents the reciprocity calibration, the results of Inmetro in recent key comparisons and the research that is being developed for the implementation of reciprocity calibration in diffuse field

  11. New calibration technique for KCD-based megavoltage imaging

    Science.gov (United States)

    Samant, Sanjiv S.; Zheng, Wei; DiBianca, Frank A.; Zeman, Herbert D.; Laughter, Joseph S.

    1999-05-01

    In megavoltage imaging, current commercial electronic portal imaging devices (EPIDs), despite having the advantage of immediate digital imaging over film, suffer from poor image contrast and spatial resolution. The feasibility of using a kinestatic charge detector (KCD) as an EPID to provide superior image contrast and spatial resolution for portal imaging has already been demonstrated in a previous paper. The KCD system had the additional advantage of requiring an extremely low dose per acquired image, allowing for superior imaging to be reconstructed form a single linac pulse per image pixel. The KCD based images utilized a dose of two orders of magnitude less that for EPIDs and film. Compared with the current commercial EPIDs and film, the prototype KCD system exhibited promising image qualities, despite being handicapped by the use of a relatively simple image calibration technique, and the performance limits of medical linacs on the maximum linac pulse frequency and energy flux per pulse delivered. This image calibration technique fixed relative image pixel values based on a linear interpolation of extrema provided by an air-water calibration, and accounted only for channel-to-channel variations. The counterpart of this for area detectors is the standard flat fielding method. A comprehensive calibration protocol has been developed. The new technique additionally corrects for geometric distortions due to variations in the scan velocity, and timing artifacts caused by mis-synchronization between the linear accelerator and the data acquisition system (DAS). The role of variations in energy flux (2 - 3%) on imaging is demonstrated to be not significant for the images considered. The methodology is presented, and the results are discussed for simulated images. It also allows for significant improvements in the signal-to- noise ratio (SNR) by increasing the dose using multiple images without having to increase the linac pulse frequency or energy flux per pulse. The

  12. Ovarian volume and antral follicle count assessed by MRI and transvaginal ultrasonography: a methodological study.

    Science.gov (United States)

    Leonhardt, Henrik; Gull, Berit; Stener-Victorin, Elisabet; Hellström, Mikael

    2014-03-01

    Ultrasonographic measurements of ovarian volume and antral follicle count are of clinical importance as diagnostic features of polycystic ovarian syndrome (PCOS), and as a parameter in estimation of ovarian follicular reserve in infertility care. To compare two-dimensional (2D)/three-dimensional (3D) transvaginal ultrasonography (TVUS) and magnetic resonance imaging (MRI) for estimation of ovarian volume and antral follicle count, and to assess reproducibility and inter-observer agreement of MRI measurements. Volumes of 172 ovaries in 99 women aged 21-37 years were calculated (length x width x height x 0.523) with conventional 2D TVUS and 2D MRI. Semi-automatic estimates of ovarian volumes were obtained by 3D MRI. Antral follicles were counted manually on 2D MRI and automatically by 3D TVUS (SonoAVC), and stratified according to follicle size. Mean ovarian volume assessed by 2D TVUS (13.1 ± 6.4 mL) was larger than assessed by 2D MRI (9.6 ± 4.1) and 3D MRI (11.4 ± 4.5) (P 0.77. 2D MRI reveals more antral follicles, especially of small size, than 3D TVUS. Ovarian volume estimation by MRI provides smaller volumes than by the reference standard 2D TVUS. Ovarian volume estimation by 3D MRI, allowing independence of non-ellipsoid ovarian shape measurement errors, provides volumes closer to 2D TVUS values than does 2D MRI. Reproducibility and inter-observer agreement of 2D MRI measurements of ovarian volume and total follicle count are good.

  13. Calibrated Properties Model

    International Nuclear Information System (INIS)

    Ahlers, C.F.; Liu, H.H.

    2001-01-01

    The purpose of this Analysis/Model Report (AMR) is to document the Calibrated Properties Model that provides calibrated parameter sets for unsaturated zone (UZ) flow and transport process models for the Yucca Mountain Site Characterization Project (YMP). This work was performed in accordance with the AMR Development Plan for U0035 Calibrated Properties Model REV00 (CRWMS M and O 1999c). These calibrated property sets include matrix and fracture parameters for the UZ Flow and Transport Model (UZ Model), drift seepage models, drift-scale and mountain-scale coupled-processes models, and Total System Performance Assessment (TSPA) models as well as Performance Assessment (PA) and other participating national laboratories and government agencies. These process models provide the necessary framework to test conceptual hypotheses of flow and transport at different scales and predict flow and transport behavior under a variety of climatic and thermal-loading conditions

  14. Radiometric Calibration of Osmi Imagery Using Solar Calibration

    Directory of Open Access Journals (Sweden)

    Dong-Han Lee

    2000-12-01

    Full Text Available OSMI (Ocean Scanning Multi-Spectral Imager raw image data (Level 0 were acquired and radiometrically corrected. We have applied two methods, using solar & dark calibration data from OSMI sensor and comparing with the SeaWiFS data, to the radiometric correction of OSMI raw image data. First, we could get the values of the gain and the offset for each pixel and each band from comparing the solar & dark calibration data with the solar input radiance values, calculated from the transmittance, BRDF (Bidirectional Reflectance Distribution Function and the solar incidence angle (¥â,¥è of OSMI sensor. Applying this calibration data to OSMI raw image data, we got the two odd results, the lower value of the radiometric corrected image data than the expected value, and the Venetian Blind Effect in the radiometric corrected image data. Second, we could get the reasonable results from comparing OSMI raw image data with the SeaWiFS data, and get a new problem of OSMI sensor.

  15. INTRODUCING NOVEL GENERATION OF HIGH ACCURACY CAMERA OPTICAL-TESTING AND CALIBRATION TEST-STANDS FEASIBLE FOR SERIES PRODUCTION OF CAMERAS

    Directory of Open Access Journals (Sweden)

    M. Nekouei Shahraki

    2015-12-01

    Full Text Available The recent advances in the field of computer-vision have opened the doors of many opportunities for taking advantage of these techniques and technologies in many fields and applications. Having a high demand for these systems in today and future vehicles implies a high production volume of video cameras. The above criterions imply that it is critical to design test systems which deliver fast and accurate calibration and optical-testing capabilities. In this paper we introduce new generation of test-stands delivering high calibration quality in single-shot calibration of fisheye surround-view cameras. This incorporates important geometric features from bundle-block calibration, delivers very high (sub-pixel calibration accuracy, makes possible a very fast calibration procedure (few seconds, and realizes autonomous calibration via machines. We have used the geometrical shape of a Spherical Helix (Type: 3D Spherical Spiral with special geometrical characteristics, having a uniform radius which corresponds to the uniform motion. This geometrical feature was mechanically realized using three dimensional truncated icosahedrons which practically allow the implementation of a spherical helix on multiple surfaces. Furthermore the test-stand enables us to perform many other important optical tests such as stray-light testing, enabling us to evaluate the certain qualities of the camera optical module.

  16. Calibration sets and the accuracy of vibrational scaling factors: A case study with the X3LYP hybrid functional

    Science.gov (United States)

    Teixeira, Filipe; Melo, André; Cordeiro, M. Natália D. S.

    2010-09-01

    A linear least-squares methodology was used to determine the vibrational scaling factors for the X3LYP density functional. Uncertainties for these scaling factors were calculated according to the method devised by Irikura et al. [J. Phys. Chem. A 109, 8430 (2005)]. The calibration set was systematically partitioned according to several of its descriptors and the scaling factors for X3LYP were recalculated for each subset. The results show that the scaling factors are only significant up to the second digit, irrespective of the calibration set used. Furthermore, multivariate statistical analysis allowed us to conclude that the scaling factors and the associated uncertainties are independent of the size of the calibration set and strongly suggest the practical impossibility of obtaining vibrational scaling factors with more than two significant digits.

  17. Calibration of the NEXT-White Detector using $^{83m}\\mathrm{Kr}$ Decays

    Energy Technology Data Exchange (ETDEWEB)

    Martínez-Lema, G.; et al.

    2018-04-05

    The NEXT-White (NEW) detector is currently the largest radio-pure high pressure gas xenon time projection chamber with electroluminescent readout in the world. NEXT-White has been operating at Laboratorio Subterr\\'aneo de Canfranc (LSC) since October 2016. This paper describes the calibrations performed with $^{83m}\\mathrm{Kr}$ decays during a long run taken from March to November 2017 (Run II). Krypton calibrations are used to correct for the finite drift-electron lifetime as well as for the dependence of the measured energy on the event position which is mainly caused by variations in solid angle coverage. After producing calibration maps to correct for both effects we measure an excellent energy resolution for 41.5 keV point-like deposits of (4.55 $\\pm$ 0.01) % FWHM in the full chamber and (3.88 $\\pm$ 0.04) % FWHM in a restricted fiducial volume. Using naive 1/$\\sqrt{E}$ scaling, these values translate into FWHM resolutions of (0.592 $\\pm$ 0.001) % FWHM and (0.504 $\\pm$ 0.005) % at the $Q_{\\beta\\beta}$ energy of xenon double beta decay (2458 keV), well within range of our target value of 1%.

  18. AFM lateral force calibration for an integrated probe using a calibration grating

    International Nuclear Information System (INIS)

    Wang, Huabin; Gee, Michelle L.

    2014-01-01

    Atomic force microscopy (AFM) friction measurements on hard and soft materials remain a challenge due to the difficulties associated with accurately calibrating the cantilever for lateral force measurement. One of the most widely accepted lateral force calibration methods is the wedge method. This method is often used in a simplified format but in so doing sacrifices accuracy. In the present work, we have further developed the wedge method to provide a lateral force calibration method for integrated AFM probes that is easy to use without compromising accuracy and reliability. Raw friction calibration data are collected from a single scan image by continuous ramping of the set point as the facets of a standard grating are scanned. These data are analysed to yield an accurate lateral force conversion/calibration factor that is not influenced by adhesion forces or load deviation. By demonstrating this new calibration method, we illustrate its reliability, speed and ease of execution. This method makes accessible reliable boundary lubrication studies on adhesive and heterogeneous surfaces that require spatial resolution of frictional forces. - Highlights: • We develop a simple and accurate method for lateral force calibration in AFM friction measurements. • We detail the basis of the method and illustrate how to use it and its reliability with example data. • Our method is easy, accurate and accounts for the affects of adhesion on friction measurements. • The method is applicable to integrated probes, as opposed to colloidal probes. • This allows accurate AFM friction measurements on spatially heterogeneous and adhesive surfaces

  19. Mass Spectrometric Calibration of Controlled Fluoroform Leak Rate Devices Technique and Uncertainty Analysis

    CERN Document Server

    Balsley, S D; Laduca, C A

    2003-01-01

    Controlled leak rate devices of fluoroform on the order of 10 sup - sup 8 atm centre dot cc sec sup - sup 1 at 25 C are used to calibrate QC-1 War Reserve neutron tube exhaust stations for leak detection sensitivity. Close-out calibration of these tritium-contaminated devices is provided by the Gas Dynamics and Mass Spectrometry Laboratory, Organization 14406, which is a tritium analytical facility. The mass spectrometric technique used for the measurement is discussed, as is the first principals calculation (pressure, volume, temperature and time). The uncertainty of the measurement is largely driven by contributing factors in the determination of P, V and T. The expanded uncertainty of the leak rate measurement is shown to be 4.42%, with a coverage factor of 3 (k=3).

  20. An improved calibration technique for measurement of 41Ar activity in gaseous effluents

    International Nuclear Information System (INIS)

    Jose, M.T.; Raghunath, V.M.

    1992-01-01

    One of the methods used generally for the calibration of a gaseous effluent monitoring system is to estimate the efficiency of the detector system with a point source of 22 Na. The uncertainty in this method is the assumption that a point source kept at a distance from the detector and a volume source nearer the detector will have the same efficiency. To overcome this problem, a technique of using 22 Na in liquid form has been standardised. The uncertainties in the strength of the liquid source is avoided by proper intercomparison of the same with a reference standard traceable to a primary standard. The NaI system has been calibrated for estimating the 41 Ar activity from the peak area counts in the MCA and SCA systems respectively. (author). 1 ref., 1 tab

  1. Potentiometric determination of ibuprofen, indomethacin and naproxen using an artificial neural network calibration

    Directory of Open Access Journals (Sweden)

    A. HAKAN AKTAŞ

    2008-01-01

    Full Text Available In this study, three anti-inflammatory agents, namely ibuprofen, indomethacin and naproxen, were titrated potentiometrically using tetrabutylammonium hydroxide in acetonitrile solvent under a nitrogen atmosphere at 25 °C. MATLAB 7.0 software was applied for data treatment as a multivariate calibration tool in the potentiometric titration procedure. An artificial neural network (ANN was used as a multivariate calibration tool in the potentiometric titration to model the complex non-linear relationship between ibuprofen, indomethacin and naproxen concentrations and the millivolt (mV of the solutions measured after the addition of different volumes of the titrant. The optimized network predicted the concentrations of agents in synthetic mixtures. The results showed that the employed ANN can precede the titration data with an average relative error of prediction of less than 2.30 %.

  2. X radiation qualities characterization following the standard IEC 61267 recommendations at the calibration laboratory of IPEN

    International Nuclear Information System (INIS)

    Franciscatto, Priscila Cerutti

    2009-01-01

    This work presents a methodology for the X radiation qualities characterization following the new recommendations of the standard 61267 of the International Electrotechnical Commission (IEC) to establish a new procedure for calibration of dosimetric systems used in the field of diagnostic radiology. The reference qualities radiation of IEC 61267: RQR 2 to RQR 10, RQA 2 to RQA 10, RQB 2 to RQB 10 and RQN 2 to RQN 10 were implanted at the calibration laboratory of IPEN (LCI). Their characteristics were analyzed through measurements of beam parameters such as: Practical peak voltage (PPV), specific additional filtrations for each qualities (high purity aluminum of about 99.9%), 1st and 2nd Half Value Layers, homogeneity coefficient. The inherent filtration of the X ray tube was also determined. With the establishment of these radiation qualities, the LCI will be ready to calibrate the measuring instruments of radiation in the new qualities, allowing an improvement in radiological services offered by IPEN. (author)

  3. Mixing volume determination in batch transfers through sonic detectors

    Energy Technology Data Exchange (ETDEWEB)

    Baptista, Renan Martins [PETROBRAS S.A., Rio de Janeiro, RJ (Brazil). Centro de Pesquisas]. E-mail: renan@cenpes.petrobras.com.br; Rachid, Felipe Bastos de Freitas [Universidade Federal Fluminense, Niteroi, RJ (Brazil). Dept. de Engenharia Mecanica]. E-mail: rachid@mec.uff.br; Araujo, Jose Henrique Carneiro de [Universidade Federal Fluminense, Niteroi, RJ (Brazil). Dept. de Ciencia da Computacao]. E-mail: jhca@dcc.ic.uff.br

    2000-07-01

    An experimental methodology to evaluate mixing volumes in batch transfers by means of sonic detectors has been reported in this paper. Mixing volumes have then been computed in a transfer of diesel/gasoline carried out through a pipeline operated by Petrobras for different interface points. It has been shown that an adequate choice of the interface points is crucial for keeping the mixing volume uncertainty within acceptable limits. (author)

  4. Methodologies and Intelligent Systems for Technology Enhanced Learning

    CERN Document Server

    Gennari, Rosella; Vittorini, Pierpaolo; Prieta, Fernando

    2015-01-01

    This volume presents recent research on Methodologies and Intelligent Systems for Technology Enhanced Learning. It contains the contributions of MIS4TEL 2015, which took place in Salamanca, Spain,. On June 3rd to 5th 2015. Like the previous edition, this proceedings and the conference is an open forum for discussing intelligent systems for Technology Enhanced Learning and empirical methodologies for their design or evaluation MIS4TEL’15 conference has been organized by University of L’aquila, Free University of Bozen-Bolzano and the University of Salamanca.  .

  5. Approaches to highly parameterized inversion-A guide to using PEST for groundwater-model calibration

    Science.gov (United States)

    Doherty, John E.; Hunt, Randall J.

    2010-01-01

    Highly parameterized groundwater models can create calibration difficulties. Regularized inversion-the combined use of large numbers of parameters with mathematical approaches for stable parameter estimation-is becoming a common approach to address these difficulties and enhance the transfer of information contained in field measurements to parameters used to model that system. Though commonly used in other industries, regularized inversion is somewhat imperfectly understood in the groundwater field. There is concern that this unfamiliarity can lead to underuse, and misuse, of the methodology. This document is constructed to facilitate the appropriate use of regularized inversion for calibrating highly parameterized groundwater models. The presentation is directed at an intermediate- to advanced-level modeler, and it focuses on the PEST software suite-a frequently used tool for highly parameterized model calibration and one that is widely supported by commercial graphical user interfaces. A brief overview of the regularized inversion approach is provided, and techniques for mathematical regularization offered by PEST are outlined, including Tikhonov, subspace, and hybrid schemes. Guidelines for applying regularized inversion techniques are presented after a logical progression of steps for building suitable PEST input. The discussion starts with use of pilot points as a parameterization device and processing/grouping observations to form multicomponent objective functions. A description of potential parameter solution methodologies and resources available through the PEST software and its supporting utility programs follows. Directing the parameter-estimation process through PEST control variables is then discussed, including guidance for monitoring and optimizing the performance of PEST. Comprehensive listings of PEST control variables, and of the roles performed by PEST utility support programs, are presented in the appendixes.

  6. Revised VESCAL: Vessel calibration data analysis program. Improvement of a model for non-linear parts of annular and slab tanks

    International Nuclear Information System (INIS)

    Yanagisawa, Hiroshi

    1995-05-01

    For the purpose of the nuclear material accountancy and control for NUCEF: the Nuclear Fuel Cycle Safety Engineering Research Facility, the vessel calibration data analysis program: VESCAL is revised, and a new model for non-linear parts of annular and slab tanks is added to the program. The new model has three unknown parameters, and liquid level is expressed as a square root function with respect to liquid volume. Using the new model, an accurate calibration function on the level and volume data for non-linear parts of annular and slab tanks can be obtained with the smaller number of unknown parameters, compared with a polynomial function model. As a result of benchmark tests for this revision, it was proved that numerical results computed with VESCAL well agreed with those by a statistical analysis program package which is widely used. In addition, the new model would be useful for carrying out data analyses on the vessel calibration at the other bulk handling facilities as well as at NUCEF. This paper describes summary of the program, computational methods and results of benchmark tests concerning this revision. (author)

  7. The GERDA calibration system

    Energy Technology Data Exchange (ETDEWEB)

    Baudis, Laura; Froborg, Francis; Tarka, Michael; Bruch, Tobias; Ferella, Alfredo [Physik-Institut, Universitaet Zuerich (Switzerland); Collaboration: GERDA-Collaboration

    2012-07-01

    A system with three identical custom made units is used for the energy calibration of the GERDA Ge diodes. To perform a calibration the {sup 228}Th sources are lowered from the parking positions at the top of the cryostat. Their positions are measured by two independent modules. One, the incremental encoder, counts the holes in the perforated steel band holding the sources, the other measures the drive shaft's angular position even if not powered. The system can be controlled remotely by a Labview program. The calibration data is analyzed by an iterative calibration algorithm determining the calibration functions for different energy reconstruction algorithms and the resolution of several peaks in the {sup 228}Th spectrum is determined. A Monte Carlo simulation using the GERDA simulation software MAGE has been performed to determine the background induced by the sources in the parking positions.

  8. The Dark Energy Survey Data Processing and Calibration System

    Energy Technology Data Exchange (ETDEWEB)

    Mohr, Joseph J. [Munich U.; Armstrong, Robert [Penn State U.; Bertin, Emmanuel [Paris, Inst. Astrophys.; Daues, Gregory E. [NCSA, Urbana; Desai, Shantanu [Munich U.; Gower, Michelle [NCSA, Urbana; Gruendl, Robert [Illinois U., Urbana (main); Hanlon, William [Illinois U., Urbana (main); Kuropatkin, Nikolay [Fermilab; Lin, Huan [Fermilab; Marriner, John [Fermilab; Petravick, Don; Sevilla, Ignacio [Madrid, CIEMAT; Swanson, Molly [Harvard-Smithsonian Ctr. Astrophys.; Tomashek, Todd [NCSA, Urbana; Tucker, Douglas [Fermilab; Yanny, Brian [Fermilab

    2012-09-24

    The Dark Energy Survey (DES) is a 5000 deg2 grizY survey reaching characteristic photometric depths of 24th magnitude (10 sigma) and enabling accurate photometry and morphology of objects ten times fainter than in SDSS. Preparations for DES have included building a dedicated 3 deg2 CCD camera (DECam), upgrading the existing CTIO Blanco 4m telescope and developing a new high performance computing (HPC) enabled data management system (DESDM). The DESDM system will be used for processing, calibrating and serving the DES data. The total data volumes are high (~2PB), and so considerable effort has gone into designing an automated processing and quality control system. Special purpose image detrending and photometric calibration codes have been developed to meet the data quality requirements, while survey astrometric calibration, coaddition and cataloging rely on new extensions of the AstrOmatic codes which now include tools for PSF modeling, PSF homogenization, PSF corrected model fitting cataloging and joint model fitting across multiple input images. The DESDM system has been deployed on dedicated development clusters and HPC systems in the US and Germany. An extensive program of testing with small rapid turn-around and larger campaign simulated datasets has been carried out. The system has also been tested on large real datasets, including Blanco Cosmology Survey data from the Mosaic2 camera. In Fall 2012 the DESDM system will be used for DECam commissioning, and, thereafter, the system will go into full science operations.

  9. RF impedance measurement calibration

    International Nuclear Information System (INIS)

    Matthews, P.J.; Song, J.J.

    1993-01-01

    The intent of this note is not to explain all of the available calibration methods in detail. Instead, we will focus on the calibration methods of interest for RF impedance coupling measurements and attempt to explain: (1). The standards and measurements necessary for the various calibration techniques. (2). The advantages and disadvantages of each technique. (3). The mathematical manipulations that need to be applied to the measured standards and devices. (4). An outline of the steps needed for writing a calibration routine that operated from a remote computer. For further details of the various techniques presented in this note, the reader should consult the references

  10. Robot-assisted automatic ultrasound calibration.

    Science.gov (United States)

    Aalamifar, Fereshteh; Cheng, Alexis; Kim, Younsu; Hu, Xiao; Zhang, Haichong K; Guo, Xiaoyu; Boctor, Emad M

    2016-10-01

    Ultrasound (US) calibration is the process of determining the unknown transformation from a coordinate frame such as the robot's tooltip to the US image frame and is a necessary task for any robotic or tracked US system. US calibration requires submillimeter-range accuracy for most applications, but it is a time-consuming and repetitive task. We provide a new framework for automatic US calibration with robot assistance and without the need for temporal calibration. US calibration based on active echo (AE) phantom was previously proposed, and its superiority over conventional cross-wire phantom-based calibration was shown. In this work, we use AE to guide the robotic arm motion through the process of data collection; we combine the capability of the AE point to localize itself in the frame of the US image with the automatic motion of the robotic arm to provide a framework for calibrating the arm to the US image automatically. We demonstrated the efficacy of the automated method compared to the manual method through experiments. To highlight the necessity of frequent ultrasound calibration, it is demonstrated that the calibration precision changed from 1.67 to 3.20 mm if the data collection is not repeated after a dismounting/mounting of the probe holder. In a large data set experiment, similar reconstruction precision of automatic and manual data collection was observed, while the time was reduced by 58 %. In addition, we compared ten automatic calibrations with ten manual ones, each performed in 15 min, and showed that all the automatic ones could converge in the case of setting the initial matrix as identity, while this was not achieved by manual data sets. Given the same initial matrix, the repeatability of the automatic was [0.46, 0.34, 0.80, 0.47] versus [0.42, 0.51, 0.98, 1.15] mm in the manual case for the US image four corners. The submillimeter accuracy requirement of US calibration makes frequent data collections unavoidable. We proposed an automated

  11. Use of lean and six sigma methodology to improve operating room efficiency in a high-volume tertiary-care academic medical center.

    Science.gov (United States)

    Cima, Robert R; Brown, Michael J; Hebl, James R; Moore, Robin; Rogers, James C; Kollengode, Anantha; Amstutz, Gwendolyn J; Weisbrod, Cheryl A; Narr, Bradly J; Deschamps, Claude

    2011-07-01

    Operating rooms (ORs) are resource-intense and costly hospital units. Maximizing OR efficiency is essential to maintaining an economically viable institution. OR efficiency projects often focus on a limited number of ORs or cases. Efforts across an entire OR suite have not been reported. Lean and Six Sigma methodologies were developed in the manufacturing industry to increase efficiency by eliminating non-value-added steps. We applied Lean and Six Sigma methodologies across an entire surgical suite to improve efficiency. A multidisciplinary surgical process improvement team constructed a value stream map of the entire surgical process from the decision for surgery to discharge. Each process step was analyzed in 3 domains, ie, personnel, information processed, and time. Multidisciplinary teams addressed 5 work streams to increase value at each step: minimizing volume variation; streamlining the preoperative process; reducing nonoperative time; eliminating redundant information; and promoting employee engagement. Process improvements were implemented sequentially in surgical specialties. Key performance metrics were collected before and after implementation. Across 3 surgical specialties, process redesign resulted in substantial improvements in on-time starts and reduction in number of cases past 5 pm. Substantial gains were achieved in nonoperative time, staff overtime, and ORs saved. These changes resulted in substantial increases in margin/OR/day. Use of Lean and Six Sigma methodologies increased OR efficiency and financial performance across an entire operating suite. Process mapping, leadership support, staff engagement, and sharing performance metrics are keys to enhancing OR efficiency. The performance gains were substantial, sustainable, positive financially, and transferrable to other specialties. Copyright © 2011 American College of Surgeons. Published by Elsevier Inc. All rights reserved.

  12. Modeling microelectrode biosensors: free-flow calibration can substantially underestimate tissue concentrations.

    Science.gov (United States)

    Newton, Adam J H; Wall, Mark J; Richardson, Magnus J E

    2017-03-01

    Microelectrode amperometric biosensors are widely used to measure concentrations of analytes in solution and tissue including acetylcholine, adenosine, glucose, and glutamate. A great deal of experimental and modeling effort has been directed at quantifying the response of the biosensors themselves; however, the influence that the macroscopic tissue environment has on biosensor response has not been subjected to the same level of scrutiny. Here we identify an important issue in the way microelectrode biosensors are calibrated that is likely to have led to underestimations of analyte tissue concentrations. Concentration in tissue is typically determined by comparing the biosensor signal to that measured in free-flow calibration conditions. In a free-flow environment the concentration of the analyte at the outer surface of the biosensor can be considered constant. However, in tissue the analyte reaches the biosensor surface by diffusion through the extracellular space. Because the enzymes in the biosensor break down the analyte, a density gradient is set up resulting in a significantly lower concentration of analyte near the biosensor surface. This effect is compounded by the diminished volume fraction (porosity) and reduction in the diffusion coefficient due to obstructions (tortuosity) in tissue. We demonstrate this effect through modeling and experimentally verify our predictions in diffusive environments. NEW & NOTEWORTHY Microelectrode biosensors are typically calibrated in a free-flow environment where the concentrations at the biosensor surface are constant. However, when in tissue, the analyte reaches the biosensor via diffusion and so analyte breakdown by the biosensor results in a concentration gradient and consequently a lower concentration around the biosensor. This effect means that naive free-flow calibration will underestimate tissue concentration. We develop mathematical models to better quantify the discrepancy between the calibration and tissue

  13. Dosimetric calibration of solid state detectors with low energy β sources

    International Nuclear Information System (INIS)

    Fidanzio, Andrea; Pia Toni, Maria; Capote, Roberto; Pena, Juan; Pasciuti, Katia; Bovi, Maurizio; Perrone, Franco; Azario, Luigi; Lazzeri, Mauro; Gaudino, Diego; Piermattei, Angelo

    2008-01-01

    A PTW Optidos plastic scintillation and a PTW natural diamond detectors were calibrated in terms of absorbed dose to water with β fields produced by 90 Sr + 90 Y and 85 Kr reference sources. Each source was characterized at the Italian National Metrological Institute - the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti of ENEA (ENEA-INMRI) - for two different series, 1 and 2, of ISO reference β-particle radiation fields. Beam flattening filters were used for the series 1 β fields to give uniform absorbed dose rates over a large area at a source-to-reference plane distance of 30 cm. The series 2 β fields were produced at source-to-reference plane distance of 10 cm, without the beam flattening filters, in order to obtain higher absorbed dose rates. The reference absorbed dose rate values were directly determined by the Italian national standard for β-particle dosimetry (a PTW extrapolation ionization chamber) for the series 1 β fields and by a calibrated transfer standard chamber, (a Capintec thin fixed-volume parallel plate ionization chamber) for the series 2 β fields. Finally the two solid state detectors were calibrated in terms of absorbed dose to water with the series 2 β field. The expanded uncertainties of the calibration coefficients obtained for the plastic scintillation dosimeter were 10% and 12% (2SD) for the 90 Sr + 90 Y and the 85 Kr sources, respectively. The expanded uncertainties obtained for the diamond dosimeter were 10% (2SD) and 16% (2SD) for the 90 Sr + 90 Y and the 85 Kr sources, respectively. The good results obtained with the 90 Sr + 90 Y and the 85 Kr β sources encourage to implement this procedure to calibrate this type of detectors at shorter distances and with other β sources of interest in brachytherapy, for example the 106 Ru source

  14. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  15. Depolarization Ratio Profiles Calibration and Observations of Aerosol and Cloud in the Tibetan Plateau Based on Polarization Raman Lidar

    Directory of Open Access Journals (Sweden)

    Guangyao Dai

    2018-03-01

    Full Text Available A brief description of the Water vapor, Cloud and Aerosol Lidar (WACAL system is provided. To calibrate the volume linear depolarization ratio, the concept of “ Δ 90 ° -calibration” is applied in this study. This effective and accurate calibration method is adjusted according to the design of WACAL. Error calculations and analysis of the gain ratio, calibrated volume linear depolarization ratio and particle linear depolarization ratio are provided as well. In this method, the influences of the gain ratio, the rotation angle of the plane of polarization and the polarizing beam splitter are discussed in depth. Two groups of measurements with half wave plate (HWP at angles of (0 ° , 45 ° and (22.5 ° , −22.5 ° are operated to calibrate the volume linear depolarization ratio. Then, the particle linear depolarization ratios measured by WACAL and CALIOP (the Cloud-Aerosol Lidar with Orthogonal Polarization during the simultaneous observations were compared. Good agreements are found. The calibration method was applied in the third Tibetan Plateau Experiment of Atmospheric Sciences (TIPEX III in 2013 and 2014 in China. Vertical profiles of the particle depolarization ratio of clouds and aerosol in the Tibetan Plateau were measured with WACAL in Litang (30.03° N, 100.28° E, 3949 m above sea level (a.s.l. in 2013 and Naqu (31.48° N, 92.06° E, 4508 m a.s.l. in 2014. Then an analysis on the polarizing properties of the aerosol, clouds and cirrus over the Tibetan Plateau is provided. The particle depolarization ratio of cirrus clouds varies from 0.36 to 0.52, with a mean value of 0.44 ± 0.04. Cirrus clouds occurred between 5.2 and 12 km above ground level (a.g.l.. The cloud thickness ranges from 0.12 to 2.55 km with a mean thickness of 1.22 ± 0.70 km. It is found that the particle depolarization ratio of cirrus clouds become larger as the height increases. However, the increase rate of the particle depolarization ratio becomes smaller as

  16. Assessment of historical leak model methodology as applied to the REDOX high-level waste tank SX-108

    International Nuclear Information System (INIS)

    JONES, T.E.

    1999-01-01

    Using the Historical Leak Model approach, the estimated leak rate (and therefore, projected leak volume) for Tank 241-SX-108 could not be reproduced using the data included in the initial document describing the leak methodology. An analysis of parameters impacting tank heat load calculations strongly suggest that the historical tank operating data lack the precision and accuracy required to estimate tank leak volumes using the Historical Leak Model methodology

  17. Design Methodology for Magnetic Field-Based Soft Tri-Axis Tactile Sensors.

    Science.gov (United States)

    Wang, Hongbo; de Boer, Greg; Kow, Junwai; Alazmani, Ali; Ghajari, Mazdak; Hewson, Robert; Culmer, Peter

    2016-08-24

    Tactile sensors are essential if robots are to safely interact with the external world and to dexterously manipulate objects. Current tactile sensors have limitations restricting their use, notably being too fragile or having limited performance. Magnetic field-based soft tactile sensors offer a potential improvement, being durable, low cost, accurate and high bandwidth, but they are relatively undeveloped because of the complexities involved in design and calibration. This paper presents a general design methodology for magnetic field-based three-axis soft tactile sensors, enabling researchers to easily develop specific tactile sensors for a variety of applications. All aspects (design, fabrication, calibration and evaluation) of the development of tri-axis soft tactile sensors are presented and discussed. A moving least square approach is used to decouple and convert the magnetic field signal to force output to eliminate non-linearity and cross-talk effects. A case study of a tactile sensor prototype, MagOne, was developed. This achieved a resolution of 1.42 mN in normal force measurement (0.71 mN in shear force), good output repeatability and has a maximum hysteresis error of 3.4%. These results outperform comparable sensors reported previously, highlighting the efficacy of our methodology for sensor design.

  18. A design of an on-orbit radiometric calibration device for high dynamic range infrared remote sensors

    Science.gov (United States)

    Sheng, Yicheng; Jin, Weiqi; Dun, Xiong; Zhou, Feng; Xiao, Si

    2017-10-01

    With the demand of quantitative remote sensing technology growing, high reliability as well as high accuracy radiometric calibration technology, especially the on-orbit radiometric calibration device has become an essential orientation in term of quantitative remote sensing technology. In recent years, global launches of remote sensing satellites are equipped with innovative on-orbit radiometric calibration devices. In order to meet the requirements of covering a very wide dynamic range and no-shielding radiometric calibration system, we designed a projection-type radiometric calibration device for high dynamic range sensors based on the Schmidt telescope system. In this internal radiometric calibration device, we select the EF-8530 light source as the calibration blackbody. EF-8530 is a high emittance Nichrome (Ni-Cr) reference source. It can operate in steady or pulsed state mode at a peak temperature of 973K. The irradiance from the source was projected to the IRFPA. The irradiance needs to ensure that the IRFPA can obtain different amplitude of the uniform irradiance through the narrow IR passbands and cover the very wide dynamic range. Combining the internal on-orbit radiometric calibration device with the specially designed adaptive radiometric calibration algorithms, an on-orbit dynamic non-uniformity correction can be accomplished without blocking the optical beam from outside the telescope. The design optimizes optics, source design, and power supply electronics for irradiance accuracy and uniformity. The internal on-orbit radiometric calibration device not only satisfies a series of indexes such as stability, accuracy, large dynamic range and uniformity of irradiance, but also has the advantages of short heating and cooling time, small volume, lightweight, low power consumption and many other features. It can realize the fast and efficient relative radiometric calibration without shielding the field of view. The device can applied to the design and

  19. A Study on Relative Radiometric Calibration without Calibration Field for YG-25

    Directory of Open Access Journals (Sweden)

    ZHANG Guo

    2017-08-01

    Full Text Available YG-25 is the first agility optical remote sensing satellite of China to acquire the sub-meter imagery of the earth. The side slither calibration technique is an on-orbit maneuver that has been used to flat-field image data acquired over the uniform calibration field. However, imaging to the single uniform calibration field cannot afford to calibrate the full dynamic response range of the sensor and reduces the efficiency. The paper proposes a new relative radiometric calibration method that a 90-degree yaw maneuver is performed over any non-uniform features of the Earth for YG-25. Meanwhile, we use an enhanced side slither image horizontal correction method based on line segment detector(LSDalgorithm to solve the side slither image over-shifted problem.The shifted results are compared with other horizontal correction method. The histogram match algorithm is used to calculate the relative gains of all detectors. The correctness and validity of the proposed method are validated by using the YG-25 on-board side slither data. The results prove that the mean streaking metrics of relative correction images of YG-25 is better 0.07%, the noticeable striping artifact and residual noise are removed, the calibration accuracy of side slither technique based on non-uniform features is superior to life image statistics of sensor's life span.

  20. Pulse-based internal calibration of polarimetric SAR

    DEFF Research Database (Denmark)

    Dall, Jørgen; Skou, Niels; Christensen, Erik Lintz

    1994-01-01

    Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops and devel......Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops...

  1. Use of calibration methodology of gamma cameras for the workers surveillance using a thyroid simulator; Uso de una metodologia de calibracion de camaras gamma para la vigilancia de trabajadores usando un simulador de tiroides

    Energy Technology Data Exchange (ETDEWEB)

    Alfaro, M.; Molina, G.; Vazquez, R.; Garcia, O., E-mail: mercedes.alfaro@inin.gob.m [ININ, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)

    2010-09-15

    In Mexico there are a significant number of nuclear medicine centers in operation. For what the accidents risk related to the transport and manipulation of open sources used in nuclear medicine can exist. The National Institute of Nuclear Research (ININ) has as objective to establish a simple and feasible methodology for the workers surveillance related with the field of the nuclear medicine. This radiological surveillance can also be applied to the public in the event of a radiological accident. To achieve this it intends to use the available equipment s in the nuclear medicine centers, together with the neck-thyroid simulators elaborated by the ININ to calibrate the gamma cameras. The gamma cameras have among their component elements that conform spectrometric systems like the employees in the evaluation of the internal incorporation for direct measurements, reason why, besides their use for diagnostic for image, they can be calibrated with anthropomorphic simulators and also with punctual sources for the quantification of the radionuclides activity distributed homogeneously in the human body, or located in specific organs. Inside the project IAEA-ARCAL-RLA/9/049-LXXVIII -Procedures harmonization of internal dosimetry- where 9 countries intervened (Argentina, Brazil, Colombia, Cuba, Chile, Mexico, Peru, Uruguay and Spain). It was developed a protocol of cameras gamma calibration for the determination in vivo of radionuclides. The protocol is the base to establish and integrated network in Latin America to attend in response to emergencies, using nuclear medicine centers of public hospitals of the region. The objective is to achieve the appropriate radiological protection of the workers, essential for the sure and acceptable radiation use, the radioactive materials and the nuclear energy. (Author)

  2. Site Calibration report

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Vesth, Allan

    The report describes site calibration measurements carried out on a site in Denmark. The measurements are carried out in accordance to Ref. [1]. The site calibration is carried out before a power performance measurement on a given turbine to clarify the influence from the terrain on the ratio...

  3. LANL MTI calibration team experience

    Science.gov (United States)

    Bender, Steven C.; Atkins, William H.; Clodius, William B.; Little, Cynthia K.; Christensen, R. Wynn

    2004-01-01

    The Multispectral Thermal Imager (MTI) was designed as an imaging radiometer with absolute calibration requirements established by Department of Energy (DOE) mission goals. Particular emphasis was given to water surface temperature retrieval using two mid wave and three long wave infrared spectral bands, the fundamental requirement was a surface temperature determination of 1K at the 68% confidence level. For the ten solar reflective bands a one-sigma radiometric performance goal of 3% was established. In order to address these technical challenges a calibration facility was constructed containing newly designed sources that were calibrated at NIST. Additionally, the design of the payload and its onboard calibration system supported post launch maintenance and update of the ground calibration. The on-orbit calibration philosophy also included vicarious techniques using ocean buoys, playas and other instrumented sites; these became increasingly important subsequent to an electrical failure which disabled the onboard calibration system. This paper offers various relevant lessons learned in the eight-year process of reducing to practice the calibration capability required by the scientific mission. The discussion presented will include observations pertinent to operational and procedural issues as well as hardware experiences; the validity of some of the initial assumptions will also be explored.

  4. Vibration transducer calibration techniques

    Science.gov (United States)

    Brinkley, D. J.

    1980-09-01

    Techniques for the calibration of vibration transducers used in the Aeronautical Quality Assurance Directorate of the British Ministry of Defence are presented. Following a review of the types of measurements necessary in the calibration of vibration transducers, the performance requirements of vibration transducers, which can be used to measure acceleration, velocity or vibration amplitude, are discussed, with particular attention given to the piezoelectric accelerometer. Techniques for the accurate measurement of sinusoidal vibration amplitude in reference-grade transducers are then considered, including the use of a position sensitive photocell and the use of a Michelson laser interferometer. Means of comparing the output of working-grade accelerometers with that of previously calibrated reference-grade devices are then outlined, with attention given to a method employing a capacitance bridge technique and a method to be used at temperatures between -50 and 200 C. Automatic calibration procedures developed to speed up the calibration process are outlined, and future possible extensions of system software are indicated.

  5. Results of monte Carlo calibrations of a low energy germanium detector

    International Nuclear Information System (INIS)

    Brettner-Messler, R.; Brettner-Messler, R.; Maringer, F.J.

    2006-01-01

    Normally, measurements of the peak efficiency of a gamma ray detector are performed with calibrated samples which are prepared to match the measured ones in all important characteristics like its volume, chemical composition and density. Another way to determine the peak efficiency is to calculate it with special monte Carlo programs. In principle the program 'Pencyl' from the source code 'P.E.N.E.L.O.P.E. 2003' can be used for peak efficiency calibration of a cylinder symmetric detector however exact data for the geometries and the materials is needed. The interpretation of the simulation results is not clear but we found a way to convert the data into values which can be compared to our measurement results. It is possible to find other simulation parameters which perform the same or better results. Further improvements can be expected by longer simulation times and more simulations in the questionable ranges of densities and filling heights. (N.C.)

  6. Energy calibration issues in nuclear resonant vibrational spectroscopy: observing small spectral shifts and making fast calibrations.

    Science.gov (United States)

    Wang, Hongxin; Yoda, Yoshitaka; Dong, Weibing; Huang, Songping D

    2013-09-01

    The conventional energy calibration for nuclear resonant vibrational spectroscopy (NRVS) is usually long. Meanwhile, taking NRVS samples out of the cryostat increases the chance of sample damage, which makes it impossible to carry out an energy calibration during one NRVS measurement. In this study, by manipulating the 14.4 keV beam through the main measurement chamber without moving out the NRVS sample, two alternative calibration procedures have been proposed and established: (i) an in situ calibration procedure, which measures the main NRVS sample at stage A and the calibration sample at stage B simultaneously, and calibrates the energies for observing extremely small spectral shifts; for example, the 0.3 meV energy shift between the 100%-(57)Fe-enriched [Fe4S4Cl4](=) and 10%-(57)Fe and 90%-(54)Fe labeled [Fe4S4Cl4](=) has been well resolved; (ii) a quick-switching energy calibration procedure, which reduces each calibration time from 3-4 h to about 30 min. Although the quick-switching calibration is not in situ, it is suitable for normal NRVS measurements.

  7. On chromatic and geometrical calibration

    DEFF Research Database (Denmark)

    Folm-Hansen, Jørgen

    1999-01-01

    The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... the correct interpolation method is described. For the chromatic issues of calibration we present the acquisition of colour and multi-spectral images, the chromatic aberrations and the various lens/camera based non-uniformities of the illumination of the image plane. It is described how the monochromatic...... to design calibration targets for both geometrical and chromatic calibration are described. We present some possible systematical errors on the detection of the objects in the calibration targets, if viewed in a non orthogonal angle, if the intensities are uneven or if the image blurring is uneven. Finally...

  8. A graphical method to evaluate spectral preprocessing in multivariate regression calibrations: example with Savitzky-Golay filters and partial least squares regression.

    Science.gov (United States)

    Delwiche, Stephen R; Reeves, James B

    2010-01-01

    In multivariate regression analysis of spectroscopy data, spectral preprocessing is often performed to reduce unwanted background information (offsets, sloped baselines) or accentuate absorption features in intrinsically overlapping bands. These procedures, also known as pretreatments, are commonly smoothing operations or derivatives. While such operations are often useful in reducing the number of latent variables of the actual decomposition and lowering residual error, they also run the risk of misleading the practitioner into accepting calibration equations that are poorly adapted to samples outside of the calibration. The current study developed a graphical method to examine this effect on partial least squares (PLS) regression calibrations of near-infrared (NIR) reflection spectra of ground wheat meal with two analytes, protein content and sodium dodecyl sulfate sedimentation (SDS) volume (an indicator of the quantity of the gluten proteins that contribute to strong doughs). These two properties were chosen because of their differing abilities to be modeled by NIR spectroscopy: excellent for protein content, fair for SDS sedimentation volume. To further demonstrate the potential pitfalls of preprocessing, an artificial component, a randomly generated value, was included in PLS regression trials. Savitzky-Golay (digital filter) smoothing, first-derivative, and second-derivative preprocess functions (5 to 25 centrally symmetric convolution points, derived from quadratic polynomials) were applied to PLS calibrations of 1 to 15 factors. The results demonstrated the danger of an over reliance on preprocessing when (1) the number of samples used in a multivariate calibration is low (<50), (2) the spectral response of the analyte is weak, and (3) the goodness of the calibration is based on the coefficient of determination (R(2)) rather than a term based on residual error. The graphical method has application to the evaluation of other preprocess functions and various

  9. Calibration of triaxial fluxgate gradiometer

    International Nuclear Information System (INIS)

    Vcelak, Jan

    2006-01-01

    The description of simple and fast calibration procedures used for double-probe triaxial fluxgate gradiometer is provided in this paper. The calibration procedure consists of three basic steps. In the first step both probes are calibrated independently in order to reach constant total field reading in every position. Both probes are numerically aligned in the second step in order that the gradient reading is zero in homogenous magnetic field. The third step consists of periodic drift calibration during measurement. The results and detailed description of each calibration step are presented and discussed in the paper. The gradiometer is finally verified during the detection of the metal object in the measuring grid

  10. Solar Cell Calibration and Measurement Techniques

    Science.gov (United States)

    Bailey, Sheila; Brinker, Dave; Curtis, Henry; Jenkins, Phillip; Scheiman, Dave

    2004-01-01

    The increasing complexity of space solar cells and the increasing international markets for both cells and arrays has resulted in workshops jointly sponsored by NASDA, ESA and NASA. These workshops are designed to obtain international agreement on standardized values for the AMO spectrum and constant, recommend laboratory measurement practices and establish a set of protocols for international comparison of laboratory measurements. A working draft of an ISO standard, WD15387, "Requirements for Measurement and Calibration Procedures for Space Solar Cells" was discussed with a focus on the scope of the document, a definition of primary standard cell, and required error analysis for all measurement techniques. Working groups addressed the issues of Air Mass Zero (AMO) solar constant and spectrum, laboratory measurement techniques, and te international round robin methodology. A summary is presented of the current state of each area and the formulation of the ISO document.

  11. Calibration of neutron moisture gauges and their ability to spatially determine soil water content in environmental studies

    International Nuclear Information System (INIS)

    Nyhan, J.W.; Martinez, J.L.; Langhorst, G.J.

    1994-10-01

    Several neutron moisture gauges were calibrated, and their ability to spatially determine soil water content was evaluated. In 1982, the midpoint of sensitivity of each neutron probe to the detection of hydrogen was determined, as well as the radius of investigation of each probe in crushed Bandelier Tuff with varying water contents. After determining the response of one of the moisture gauges to changes in soil water at the soil-air interface, a neutron transport model was successfully calibrated to predict spatial variations in soil water content. The model was then used to predict various shapes and volumes of crushed Bandelier Tuff interrogated by the neutron moisture gauge. From 1991 through 1994, six neutron moisture gauges were calibrated for soil water determinations in a local topsoil and crushed Bandelier Tuff, as well as for a sample of fine sand and soils from a field experiment at Hill Air Force Base. Statistical analysis of the calibration results is presented and summarized, and a final summary of practical implications for future neutron moisture gauge studies at Los Alamos is included

  12. Calibrating Legal Judgments

    Directory of Open Access Journals (Sweden)

    Frederick Schauer

    2017-09-01

    Full Text Available Objective to study the notion and essence of legal judgments calibration the possibilities of using it in the lawenforcement activity to explore the expenses and advantages of using it. Methods dialectic approach to the cognition of social phenomena which enables to analyze them in historical development and functioning in the context of the integrity of objective and subjective factors it determined the choice of the following research methods formallegal comparative legal sociological methods of cognitive psychology and philosophy. Results In ordinary life people who assess other peoplersaquos judgments typically take into account the other judgments of those they are assessing in order to calibrate the judgment presently being assessed. The restaurant and hotel rating website TripAdvisor is exemplary because it facilitates calibration by providing access to a raterrsaquos previous ratings. Such information allows a user to see whether a particular rating comes from a rater who is enthusiastic about every place she patronizes or instead from someone who is incessantly hard to please. And even when less systematized as in assessing a letter of recommendation or college transcript calibration by recourse to the decisional history of those whose judgments are being assessed is ubiquitous. Yet despite the ubiquity and utility of such calibration the legal system seems perversely to reject it. Appellate courts do not openly adjust their standard of review based on the previous judgments of the judge whose decision they are reviewing nor do judges in reviewing legislative or administrative decisions magistrates in evaluating search warrant representations or jurors in assessing witness perception. In most legal domains calibration by reference to the prior decisions of the reviewee is invisible either because it does not exist or because reviewing bodies are unwilling to admit using what they in fact know and employ. Scientific novelty for the first

  13. Standard-target calibration of an acoustic backscatter system

    Science.gov (United States)

    Foote, Kenneth G.; Martini, Marinna A.

    2010-01-01

    The standard-target method used to calibrate scientific echo sounders and other scientific sonars by a single, solid elastic sphere is being adapted to acoustic backscatter (ABS) systems. Its first application, to the AQUAscat 1000, is described. The on-axis sensitivity and directional properties of transducer beams at three operating frequencies, nominally 1, 2.5, and 4 MHz, have been determined using a 10-mm-diameter sphere of tungsten carbide with 6% cobalt binder. Preliminary results are reported for the 1-MHz transducer. Their application to measurements of suspended sediment made in situ with the same device is described. This will enable the data to be expressed directly in physical units of volume backscattering.

  14. Methodology for Gamma cameras calibration for I-131 uptake quantification in Hyperthyroidism diseases

    International Nuclear Information System (INIS)

    Lopez Diaz, A.; Palau San Pedro, A.; Martin Escuela, J. M.; Reynosa Montejo, R.; Castillo, J.; Torres Aroche, L.

    2015-01-01

    Optimization and verification of Patient-Specific Treatment Planning with unsealed I-131 sources is a desirable goal from medical and radiation protection point of view. To obtain a practical protocol to combine the estimation of the related parameters with patient's specific treatment dose in hyperthyroidism disease, 3 equipment were studied (Iodine Probe, a Philips Forte Camera with pin-hole collimators and a Mediso Nucline with HEGP for planar and SPECT techniques) and crossed calibrated. The linear behaviour on diagnostic and therapeutic activity range was verified, showing a linear correlation fitting factor R 2 > 0.99. The differences between thyroid uptake determinations in all equipment were less than 6% for therapeutic activities and less than 1.1% in the diagnostic range. The combined protocol to calculate, with only one administration of I 131 , all the necessary parameters to the treatment dose estimation in 2D or 3D, avoiding wasting time with gamma cameras, was established and verified. Following this protocol the difference between apparent and calculated activities were less than 3%. (Author)

  15. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  16. An alternative method for calibration of narrow band radiometer using a radiative transfer model

    Energy Technology Data Exchange (ETDEWEB)

    Salvador, J; Wolfram, E; D' Elia, R [Centro de Investigaciones en Laseres y Aplicaciones, CEILAP (CITEFA-CONICET), Juan B. de La Salle 4397 (B1603ALO), Villa Martelli, Buenos Aires (Argentina); Zamorano, F; Casiccia, C [Laboratorio de Ozono y Radiacion UV, Universidad de Magallanes, Punta Arenas (Chile) (Chile); Rosales, A [Universidad Nacional de la Patagonia San Juan Bosco, UNPSJB, Facultad de Ingenieria, Trelew (Argentina) (Argentina); Quel, E, E-mail: jsalvador@citefa.gov.ar [Universidad Nacional de la Patagonia Austral, Unidad Academica Rio Gallegos Avda. Lisandro de la Torre 1070 ciudad de Rio Gallegos-Sta Cruz (Argentina) (Argentina)

    2011-01-01

    The continual monitoring of solar UV radiation is one of the major objectives proposed by many atmosphere research groups. The purpose of this task is to determine the status and degree of progress over time of the anthropogenic composition perturbation of the atmosphere. Such changes affect the intensity of the UV solar radiation transmitted through the atmosphere that then interacts with living organisms and all materials, causing serious consequences in terms of human health and durability of materials that interact with this radiation. One of the many challenges that need to be faced to perform these measurements correctly is the maintenance of periodic calibrations of these instruments. Otherwise, damage caused by the UV radiation received will render any one calibration useless after the passage of some time. This requirement makes the usage of these instruments unattractive, and the lack of frequent calibration may lead to the loss of large amounts of acquired data. Motivated by this need to maintain calibration or, at least, know the degree of stability of instrumental behavior, we have developed a calibration methodology that uses the potential of radiative transfer models to model solar radiation with 5% accuracy or better relative to actual conditions. Voltage values in each radiometer channel involved in the calibration process are carefully selected from clear sky data. Thus, tables are constructed with voltage values corresponding to various atmospheric conditions for a given solar zenith angle. Then we model with a radiative transfer model using the same conditions as for the measurements to assemble sets of values for each zenith angle. The ratio of each group (measured and modeled) allows us to calculate the calibration coefficient value as a function of zenith angle as well as the cosine response presented by the radiometer. The calibration results obtained by this method were compared with those obtained with a Brewer MKIII SN 80 located in the

  17. Fitness analysis method for magnesium in drinking water with atomic absorption using quadratic curve calibration

    Directory of Open Access Journals (Sweden)

    Esteban Pérez-López

    2014-11-01

    Full Text Available Because of the importance of quantitative chemical analysis in research, quality control, sales of services and other areas of interest , and the limiting of some instrumental analysis methods for quantification with linear calibration curve, sometimes because the short linear dynamic ranges of the analyte, and sometimes by limiting the technique itself, is that there is a need to investigate a little more about the convenience of using quadratic curves for analytical quantification, which seeks demonstrate that it is a valid calculation model for chemical analysis instruments. To this was taken as an analysis method based on the technique and atomic absorption spectroscopy in particular a determination of magnesium in a sample of drinking water Tacares sector Northern Grecia, employing a nonlinear calibration curve and a curve specific quadratic behavior, which was compared with the test results obtained for the same analysis with a linear calibration curve. The results show that the methodology is valid for the determination referred to, with all confidence, since the concentrations are very similar, and as used hypothesis testing can be considered equal.

  18. A Review of Sensor Calibration Monitoring for Calibration Interval Extension in Nuclear Power Plants

    Energy Technology Data Exchange (ETDEWEB)

    Coble, Jamie B.; Meyer, Ryan M.; Ramuhalli, Pradeep; Bond, Leonard J.; Hashemian, Hash; Shumaker, Brent; Cummins, Dara

    2012-08-31

    Currently in the United States, periodic sensor recalibration is required for all safety-related sensors, typically occurring at every refueling outage, and it has emerged as a critical path item for shortening outage duration in some plants. Online monitoring can be employed to identify those sensors that require calibration, allowing for calibration of only those sensors that need it. International application of calibration monitoring, such as at the Sizewell B plant in United Kingdom, has shown that sensors may operate for eight years, or longer, within calibration tolerances. This issue is expected to also be important as the United States looks to the next generation of reactor designs (such as small modular reactors and advanced concepts), given the anticipated longer refueling cycles, proposed advanced sensors, and digital instrumentation and control systems. The U.S. Nuclear Regulatory Commission (NRC) accepted the general concept of online monitoring for sensor calibration monitoring in 2000, but no U.S. plants have been granted the necessary license amendment to apply it. This report presents a state-of-the-art assessment of online calibration monitoring in the nuclear power industry, including sensors, calibration practice, and online monitoring algorithms. This assessment identifies key research needs and gaps that prohibit integration of the NRC-approved online calibration monitoring system in the U.S. nuclear industry. Several needs are identified, including the quantification of uncertainty in online calibration assessment; accurate determination of calibration acceptance criteria and quantification of the effect of acceptance criteria variability on system performance; and assessment of the feasibility of using virtual sensor estimates to replace identified faulty sensors in order to extend operation to the next convenient maintenance opportunity. Understanding the degradation of sensors and the impact of this degradation on signals is key to

  19. Optimal Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard; Kroon, I. B.; Faber, Michael Havbro

    1994-01-01

    Calibration of partial safety factors is considered in general, including classes of structures where no code exists beforehand. The partial safety factors are determined such that the difference between the reliability for the different structures in the class considered and a target reliability...... level is minimized. Code calibration on a decision theoretical basis is also considered and it is shown how target reliability indices can be calibrated. Results from code calibration for rubble mound breakwater designs are shown....

  20. Observation models in radiocarbon calibration

    International Nuclear Information System (INIS)

    Jones, M.D.; Nicholls, G.K.

    2001-01-01

    The observation model underlying any calibration process dictates the precise mathematical details of the calibration calculations. Accordingly it is important that an appropriate observation model is used. Here this is illustrated with reference to the use of reservoir offsets where the standard calibration approach is based on a different model to that which the practitioners clearly believe is being applied. This sort of error can give rise to significantly erroneous calibration results. (author). 12 refs., 1 fig

  1. Parameters of calibration of the measurement system of 222 Rn based in LR-115

    International Nuclear Information System (INIS)

    Garcia, M.L.; Mireles, F.; Quirino, L.; Davila, I.; Lugo, F.; Pinedo, J.L.; Chavez, A.

    2003-01-01

    Since the SSNTD technique (Solid State Nuclear Track Detection) it was discovered it has been used as passive method for the detection of subnuclear particles in great variety of fields of the science. The use of the technique in measurements of 222 Rn in air have already been established implying better methodologies in the exhibition to the environment until their engraving and reading processes. The SSNTD technique is since a method by comparison since the material it can be used a single time, therefore it requires of calibration in one controlled radon atmosphere, using gauged standards. The objective of this work is to show the calibration of the devices used as radon monitors based on SSNTD. The material used as SSNTD is LR-115 Il. The standardization of the parameters used in the exhibition to radon in air, engraving and reading process, its are based on the response of the LR-115 Il, the one arrangement of the device, engraving speed and mainly the calibration factor. They are considered two types of monitors: Open camera and Closed camera, the difference among the calibration factors of both cameras is the percentage of the descendants of radon in the open camera. The standardized parameters are operation voltage of the counting system; temperature, time and concentration of the engraving solution; and thickness. (Author)

  2. Using gas blow methods to realize accurate volume measurement of radioactivity liquid

    International Nuclear Information System (INIS)

    Zhang Caiyun

    2010-01-01

    For liquid which has radioactivity, Realized the accurate volume measurement uncertainty less than 0.2% (k=2) by means of gas blow methods presented in the 'American National Standard-Nuclear Material Control-Volume Calibration Methods(ANSI N15.19-1989)' and the 'ISO Committee Drafts (ISO/TC/85/SC 5N 282 )' and Explored a set methods of Data Processing. In the article, the major problems is to solve data acquisition and function foundation and measurement uncertainty estimate. (authors)

  3. Calibration simulation. A calibration Monte-Carlo program for the OPAL jet chamber

    International Nuclear Information System (INIS)

    Biebel, O.

    1989-12-01

    A calibration Monte Carlo program has been developed as a tool to investigate the interdependence of track reconstruction and calibration constants. Three categories of calibration effects have been considered: The precise knowledge of sense wire positions, necessary to reconstruct the particle trajectories in the jet chamber. Included are the staggering and the sag of the sense wires as well as tilts and rotations of their support structures. The various contributions to the measured drift time, with special emphasis on the aberration due to the track angle and the presence of a transverse magnetic field. A very precise knowledge of the drift velocity and the Lorentz angle of the drift paths with respect to the drift field is also required. The effects degrading particle identification via energy loss dE/dx. Impurities of the gas mixture and saturation effects depending on the track angle as well as the influence of the pulse shaping-electronics have been studied. These effects have been parametrised with coefficients corresponding to the calibration constants required for track reconstruction. Excellent agreement with the input data has been achieved when determining calibration constants from Monte Carlo data generated with these parametrisations. (orig.) [de

  4. Infering and Calibrating Triadic Closure in a Dynamic Network

    Science.gov (United States)

    Mantzaris, Alexander V.; Higham, Desmond J.

    In the social sciences, the hypothesis of triadic closure contends that new links in a social contact network arise preferentially between those who currently share neighbours. Here, in a proof-of-principle study, we show how to calibrate a recently proposed evolving network model to time-dependent connectivity data. The probabilistic edge birth rate in the model contains a triadic closure term, so we are also able to assess statistically the evidence for this effect. The approach is shown to work on data generated synthetically from the model. We then apply this methodology to some real, large-scale data that records the build up of connections in a business-related social networking site, and find evidence for triadic closure.

  5. Portable compact multifunction IR calibrator

    International Nuclear Information System (INIS)

    Wyatt, C.L.; Jacobsen, L.; Steed, A.

    1988-01-01

    A compact portable multifunction calibrator designed for future sensor systems is described which enables a linearity calibration for all detectors simultaneously using a near small-area source, a high-resolution mapping of the focal plane with 10 microrad setability and with a blur of less than 100 microrad, system spectral response calibration (radiometer) using a Michelson interferometer source, relative spectral response (spectrometer) using high-temperature external commercial blackbody simulators, and an absolute calibration using an internal low-temperature extended-area source. 5 references

  6. Physiotherapy ultrasound calibrations

    International Nuclear Information System (INIS)

    Gledhill, M.

    1996-01-01

    Calibration of physiotherapy ultrasound equipment has long been a problem. Numerous surveys around the world over the past 20 years have all found that only a low percentage of the units tested had an output within 30% of that indicatd. In New Zealand, a survey carried out by the NRL in 1985 found that only 24% had an output, at the maximum setting, within + or - 20% of that indicated. The present performance Standard for new equipment (NZS 3200.2.5:1992) requires that the measured output should not deviate from that indicated by more than + or - 30 %. This may be tightened to + or - 20% in the next few years. Any calibration is only as good as the calibration equipment. Some force balances can be tested with small weights to simulate the force exerted by an ultrasound beam, but with others this is not possible. For such balances, testing may only be feasible with a calibrated source which could be used like a transfer standard. (author). 4 refs., 3 figs

  7. A special mini-extrapolation chamber for calibration of 90Sr+90Y sources

    International Nuclear Information System (INIS)

    Oliveira, Mercia L; Caldas, Linda V E

    2005-01-01

    90 Sr+ 90 Y applicators are commonly utilized in brachytherapy, including ophthalmic procedures. The recommended instruments for the calibration of these applicators are extrapolation chambers, which are ionization chambers that allow the variation of their sensitive volume. Using the extrapolation method, the absorbed dose rate at the applicator surface can be determined. The aim of the present work was to develop a mini-extrapolation chamber for the calibration of 90 Sr+ 90 Y beta ray applicators. The developed mini-chamber has a 3.0 cm outer diameter and is 11.3 cm in length. An aluminized polyester foil is used as the entrance window while the collecting electrode is made of graphited polymethylmethacrylate. This mini-chamber was tested in 90 Sr+ 90 Y radiation beams from a beta particle check source and with a plane ophthalmic applicator, showing adequate results

  8. Fitness of the analysis method of magnesium in drinking water using atomic absorption with quadratic calibration curve

    International Nuclear Information System (INIS)

    Perez-Lopez, Esteban

    2014-01-01

    The quantitative chemical analysis has been importance in research. Also, aspects like: quality control, sales of services and other areas of interest. Some instrumental analysis methods for quantification with linear calibration curve have presented limitations, because the short liner dynamic ranges of the analyte, or sometimes, by limiting the technique itself. The need has been to investigate a little more about the convenience of using quadratic calibration curves for analytical quantification, with which it has seeked demonstrate that has been a valid calculation model for chemical analysis instruments. An analysis base method is used on the technique of atomic absorption spectroscopy and in particular a determination of magnesium in a drinking water sample of the Tacares sector North of Grecia. A nonlinear calibration curve was used and specifically a curve with quadratic behavior. The same was compared with the test results obtained for the equal analysis with a linear calibration curve. The results have showed that the methodology has been valid for the determination referred with all confidence, since the concentrations have been very similar and, according to the used hypothesis testing, can be considered equal. (author) [es

  9. Soucreless efficiency calibration for HPGe detector based on medical images

    International Nuclear Information System (INIS)

    Chen Chaobin; She Ruogu; Xiao Gang; Zuo Li

    2012-01-01

    Digital phantom of patient and region of interest (supposed to be filled with isotropy volume source) are built from medical CT images. They are used to calculate the detection efficiency of HPGe detectors located outside of human body by sourceless calibration method based on a fast integral technique and MCNP code respectively, and the results from two codes are in good accord besides a max difference about 5% at intermediate energy region. The software produced in this work are in better behavior than Monte Carlo code not only in time consume but also in complexity of problem to solve. (authors)

  10. Development methodology for industrial diesel engines; Entwicklungsmethode fuer Industrie-Dieselmotoren

    Energy Technology Data Exchange (ETDEWEB)

    Bergmann, Dirk; Kech, Johannes [MTU Friedrichshafen GmbH (Germany)

    2011-11-15

    In order to remain cost-effective with relatively low production volumes in spite of the high requirements regarding emissions and durability, MTU uses a clearly structured development methodology with a close interlinking of technology and product development in the development of its large engines. For the new engine of the 4000 Series with cooled EGR, MTU applied this methodology in order to implement the emissions concept from the initial idea right through to the serial product. (orig.)

  11. Development of portable flow calibrator

    International Nuclear Information System (INIS)

    Akiyama, Kiyomitsu; Iijima, Nobuo

    1995-01-01

    In the nuclear facilities, air sniffer system is often utilized to evaluate atmospheric concentration of radioactivity in the working environment. The system collects airborne dust on the filter during some sampling period. In this method, total air flow during the sampling period is an important parameter to evaluate the radioactivity concentration correctly. Therefore, calibration for the flow meter of air sniffer system must be done periodically according to Japan Industry Standards (JIS). As we have had to available device to calibrate the flow meter in the working area, we had to remove the flow meters from the installed place and carry them to another place where calibration can be made. This work required a great deal of labor. Now we have developed a portable flow calibrator for air sniffer system which enables us to make in-site calibration of the flow meter in the working area more easily. This report describes the outline of portable flow calibrator and it's experimental results. (author)

  12. Validation of response simulation methodology of Albedo dosemeter

    International Nuclear Information System (INIS)

    Freitas, B.M.; Silva, A.X. da

    2016-01-01

    The Instituto de Radioprotecao e Dosimetria developed and runs a neutron TLD albedo individual monitoring service. To optimize the dose calculation algorithm and to infer new calibration factors, the response of this dosemeter was simulated. In order to validate this employed methodology, it was applied in the simulation of the problem of the QUADOS (Quality Assurance of Computational Tools for Dosimetry) intercomparison, aimed to evaluate dosimetric problems, one being to calculate the response of a generic albedo dosemeter. The obtained results were compared with those of other modeling and the reference one, with good agreements. (author)

  13. Laser Tracker Calibration - Testing the Angle Measurement System -

    Energy Technology Data Exchange (ETDEWEB)

    Gassner, Georg; Ruland, Robert; /SLAC

    2008-12-05

    Physics experiments at the SLAC National Accelerator Laboratory (SLAC) usually require high accuracy positioning, e. g. 100 {micro}m over a distance of 150 m or 25 {micro}m in a 10 x 10 x 3 meter volume. Laser tracker measurement systems have become one of the most important tools for achieving these accuracies when mapping components. The accuracy of these measurements is related to the manufacturing tolerances of various individual components, the resolutions of measurement systems, the overall precision of the assembly, and how well imperfections can be modeled. As with theodolites and total stations, one can remove the effects of most assembly and calibration errors by measuring targets in both direct and reverse positions and computing the mean to obtain the result. However, this approach does not compensate for errors originating from the encoder system. In order to improve and gain a better understanding of laser tracker angle measurement tolerances we extended our laboratory's capabilities with the addition of a horizontal angle calibration test stand. This setup is based on the use of a high precision rotary table providing an angular accuracy of better than 0.2 arcsec. Presently, our setup permits only tests of the horizontal angle measurement system. A test stand for vertical angle calibration is under construction. Distance measurements (LECOCQ & FUSS, 2000) are compared to an interferometer bench for distances of up to 32 m. Together both tests provide a better understanding of the instrument and how it should be operated. The observations also provide a reasonable estimate of covariance information of the measurements according to their actual performance for network adjustments.

  14. Assessing the quality of the volume-outcome relationship in uro-oncology.

    Science.gov (United States)

    Mayer, Erik K; Purkayastha, Sanjay; Athanasiou, Thanos; Darzi, Ara; Vale, Justin A

    2009-02-01

    To assess systematically the quality of evidence for the volume-outcome relationship in uro-oncology, and thus facilitate the formulating of health policy within this speciality, as 'Implementation of Improving Outcome Guidance' has led to centralization of uro-oncology based on published studies that have supported a 'higher volume-better outcome' relationship, but improved awareness of methodological drawbacks in health service research has questioned the strength of this proposed volume-outcome relationship. We systematically searched previous relevant reports and extracted all articles from 1980 onwards assessing the volume-outcome relationship for cystectomy, prostatectomy and nephrectomy at the institution and/or surgeon level. Studies were assessed for their methodological quality using a previously validated rating system. Where possible, meta-analytical methods were used to calculate overall differences in outcome measures between low and high volume healthcare providers. In all, 22 studies were included in the final analysis; 19 of these were published in the last 5 years. Only four studies appropriately explored the effect of both the institution and surgeon volume on outcome measures. Mortality and length of stay were the most frequently measured outcomes. The median total quality scores within each of the operation types were 8.5, 9 and 8 for cystectomy, prostatectomy and nephrectomy, respectively (possible maximum score 18). Random-effects modelling showed a higher risk of mortality in low-volume institutions than in higher-volume institutions for both cystectomy and nephrectomy (odds ratio 1.88, 95% confidence interval 1.54-2.29, and 1.28, 1.10-1.49, respectively). The methodological quality of volume-outcome research as applied to cystectomy, prostatectomy and nephrectomy is only modest at best. Accepting several limitations, pooled analysis confirms a higher-volume, lower-mortality relationship for cystectomy and nephrectomy. Future research should

  15. Improving process methodology for measuring plutonium burden in human urine using fission track analysis

    International Nuclear Information System (INIS)

    Krahenbuhl, M.P.; Slaughter, D.M.

    1998-01-01

    The aim of this paper is to clearly define the chemical and nuclear principles governing Fission Track Analysis (FTA) to determine environmental levels of 239 Pu in urine. The paper also addresses deficiencies in FTA methodology and introduces improvements to make FTA a more reliable research tool. Our refined methodology, described herein, includes a chemically-induced precipitation phase, followed by anion exchange chromatography and employs a chemical tracer, 236 Pu. We have been able to establish an inverse correlation between Pu recovery and sample volume and our data confirms that increases in sample volume do not result in higher accuracy or lower detection limits. We conclude that in subsequent studies, samples should be limited to approximately two liters. The Pu detection limit for a sample of this volume is 2.8 μBq/l. (author)

  16. NIST high-dose calibration services

    International Nuclear Information System (INIS)

    Humphreys, J.C.

    1989-01-01

    There is a need for the standardization of high-dose measurements used in the radiation-processing industry in order to provide assured traceability to national standards. NIST provides dosimetry calibration services to this industry. One of these services involves administration of known absorbed doses of gamma rays to customer-supplied dosimeters. The dosimeters are packaged to provide electron equilibrium conditions and are irradiated in a standard 60 Co calibration facility; this provides a calibration of that batch of dosimeters. Another service consists of supplying to a customer calibrated transfer dosimeters for irradiation with the customer's radiation source. The irradiated transfer dosimeters are then returned to NIST for analysis; the results are reported to the customer, providing a calibration of the dose rate of the customer's source. (orig.)

  17. TIME CALIBRATED OSCILLOSCOPE SWEEP CIRCUIT

    Science.gov (United States)

    Smith, V.L.; Carstensen, H.K.

    1959-11-24

    An improved time calibrated sweep circuit is presented, which extends the range of usefulness of conventional oscilloscopes as utilized for time calibrated display applications in accordance with U. S. Patent No. 2,832,002. Principal novelty resides in the provision of a pair of separate signal paths, each of which is phase and amplitude adjustable, to connect a high-frequency calibration oscillator to the output of a sawtooth generator also connected to the respective horizontal deflection plates of an oscilloscope cathode ray tube. The amplitude and phase of the calibration oscillator signals in the two signal paths are adjusted to balance out feedthrough currents capacitively coupled at high frequencies of the calibration oscillator from each horizontal deflection plate to the vertical plates of the cathode ray tube.

  18. Electron Density Calibration for Radiotherapy Treatment Planning

    International Nuclear Information System (INIS)

    Herrera-Martinez, F.; Rodriguez-Villafuerte, M.; Martinez-Davalos, A.; Ruiz-Trejo, C.; Celis-Lopez, M. A.; Larraga-Gutierrez, J. M.; Garcia-Garduno, A.

    2006-01-01

    Computed tomography (CT) images are used as basic input data for most modern radiosurgery treatment planning systems (TPS). CT data not only provide anatomic information to delineate target volumes, but also allow the introduction of corrections for tissue inhomogeneities into dose calculations during the treatment planning procedure. These corrections involve the determination of a relationship between tissue electron density (ρe) and their corresponding Hounsfield Units (HU). In this work, an elemental analysis of different commercial tissue equivalent materials using Scanning Electron Microscopy was carried out to characterize their chemical composition. The tissue equivalent materials were chosen to ensure a large range of ρe to be included in the CT scanner calibration. A phantom was designed and constructed with these materials to simulate the size of a human head

  19. Methodology for characterisation of glass fibre composite architecture

    DEFF Research Database (Denmark)

    Hansen, Jens Zangenberg; Larsen, J.B.; Østergaard, R.C.

    2012-01-01

    of the fibres. The information is used for different analyses to investigate and characterise the fibre architecture. As an example, the methodology is applied to glass fibre reinforced composites with varying fibre contents. The different fibre volume fractions (FVFs) affect the number of contact points per......The present study outlines a methodology for microstructural characterisation of fibre reinforced composites containing circular fibres. Digital micrographs of polished cross-sections are used as input to a numerical image processing tool that determines spatial mapping and radii detection...... fibre, the communal fibre distance and the local FVF. The fibre diameter distribution and packing pattern remain somewhat similar for the considered materials. The methodology is a step towards a better understanding of the composite microstructure and can be used to evaluate the interconnection between...

  20. Lidar to lidar calibration

    DEFF Research Database (Denmark)

    Fernandez Garcia, Sergio; Villanueva, Héctor

    This report presents the result of the lidar to lidar calibration performed for ground-based lidar. Calibration is here understood as the establishment of a relation between the reference lidar wind speed measurements with measurement uncertainties provided by measurement standard and corresponding...... lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from the reference lidar measurements are given for information only....