WorldWideScience

Sample records for valve thrombogenicity related

  1. Are anticoagulant independent mechanical valves within reach-fast prototype fabrication and in vitro testing of innovative bi-leaflet valve models.

    Science.gov (United States)

    Scotten, Lawrence N; Siegel, Rolland

    2015-08-01

    Exploration for causes of prosthetic valve thrombogenicity has frequently focused on forward or post-closure flow detail. In prior laboratory studies, we uncovered high amplitude flow velocities of short duration close to valve closure implying potential for substantial shear stress with subsequent initiation of blood coagulation pathways. This may be relevant to widely accepted clinical disparity between mechanical and tissue valves vis-à-vis thrombogenicity. With a series of prototype bi-leaflet mechanical valves, we attempt reduction of closure related velocities with the objective of identifying a prototype valve with thrombogenic potential similar to our tissue valve control. This iterative design approach may find application in preclinical assessment of valves for anticoagulation independence. Tested valves included: prototype mechanical bi-leaflet BVs (n=56), controls (n=2) and patented early prototype mechanicals (n=2) from other investigators. Pulsatile and quasi-steady flow systems were used for testing. Projected dynamic valve area (PDVA) was measured using previously described novel technology. Flow velocity over the open and closing periods was determined by volumetric flow rate/PDVA. For the closed valve interval, use was made of data obtained from quasi-steady back pressure/flow tests. Performance was ranked by a proposed thrombogenicity potential index (TPI) relative to tissue and mechanical control valves. Optimization of the prototype valve designs lead to a 3-D printed model (BV3D). For the mitral/aortic site, BV3D has lower TPI (1.10/1.47) relative to the control mechanical valve (3.44/3.93) and similar to the control tissue valve (ideal TPI ≤1.0). Using unique technology, rapid prototyping and thrombogenicity ranking, optimization of experimental valves for reduced thrombogenic potential was expedited and simplified. Innovative mechanical valve configurations were identified that merit consideration for further development which may bring

  2. Qualification of safety-related valve actuators

    International Nuclear Information System (INIS)

    Anon.

    1981-01-01

    This Standard describes the qualification of all types of power-driven valve actuators, including damper actuators, for safety-related functions in nuclear power generating stations. It may also be used to separately qualify actuator components. This Standard establishes the minimum requirements for, and guidance regarding, the methods and procedures for qualification of all safety-related functions of power-driven valve actuators

  3. Force measuring valve assemblies, systems including such valve assemblies and related methods

    Science.gov (United States)

    DeWall, Kevin George [Pocatello, ID; Garcia, Humberto Enrique [Idaho Falls, ID; McKellar, Michael George [Idaho Falls, ID

    2012-04-17

    Methods of evaluating a fluid condition may include stroking a valve member and measuring a force acting on the valve member during the stroke. Methods of evaluating a fluid condition may include measuring a force acting on a valve member in the presence of fluid flow over a period of time and evaluating at least one of the frequency of changes in the measured force over the period of time and the magnitude of the changes in the measured force over the period of time to identify the presence of an anomaly in a fluid flow and, optionally, its estimated location. Methods of evaluating a valve condition may include directing a fluid flow through a valve while stroking a valve member, measuring a force acting on the valve member during the stroke, and comparing the measured force to a reference force. Valve assemblies and related systems are also disclosed.

  4. Seasonal variation in thrombogenicity of blood: a word of caution.

    Science.gov (United States)

    Narang, Sumit; Banerjee, Amit; Satsangi, Deepak K; Geelani, Mohammad A

    2009-01-01

    Thrombogenicity of blood is known to have seasonal variations. The clinical implications of seasonal variations in the anticoagulation profile of patients with mechanical valves was assessed. Data of patients implanted with a mechanical heart valve for more than 3 months were collected at follow-up or on presentation to the emergency department. The mean time from the previous follow-up examination was 3.6 +/- 0.3 months. The number of patients with an international normalized ratio >3.5 and the incidence of hemorrhagic events peaked in hottest part of the year (June-July), with 128 cases of prolonged clotting and 43 hemorrhagic events in this period. The number of patients with rapid clotting and the incidence of embolic events peaked in coldest part of the year (December-January), with 120 cases of international normalized ratio <1.5 and 37 embolic events in this period. There was a significant correlation between temperature and thrombogenicity in patients with prosthetic heart valves on long-term anticoagulation.

  5. MECHANICAL HEART-VALVE PROSTHESES - SOUND LEVEL AND RELATED COMPLAINTS

    NARCIS (Netherlands)

    LAURENS, RRP; WIT, HP; EBELS, T

    In a randomised study, we investigated the sound production of mechanical heart valve prostheses and the complaints related to this sound. The CarboMedics, Bjork-Shiley monostrut and StJude Medical prostheses were compared. A-weighted levels of the pulse-like sound produced by the prosthesis were

  6. Pannus-related prosthetic valve dysfunction. Case report.

    Science.gov (United States)

    Moldovan, Maria-Sînziana; Bedeleanu, Daniela; Kovacs, Emese; Ciumărnean, Lorena; Molnar, Adrian

    2016-01-01

    Pannus-related prosthetic valve dysfunction, a complication of mechanical prosthetic valve replacement, is rare, with a slowly progressive evolution, but it can be acute, severe, requiring surgical reintervention. We present the case of a patient with a mechanical single disc aortic prosthesis, with moderate prosthesis-patient mismatch, minor pannus found on previous ultrasound examinations, who presented to our service with angina pain with a duration of 1 hour, subsequently interpreted as non-ST segment elevation myocardial infarction (NSTEMI) syndrome. Coronarography showed normal epicardial coronary arteries, an ample movement of the prosthetic disc, without evidence of coronary thromboembolism, and Gated Single-Photon Emission Computerized Tomography (SPECT) with Technetium (Tc)-99m detected no perfusion defects. Transthoracic echocardiography (TTE) evidenced a dysfunctional prosthesis due to a subvalvular mass; transesophageal echocardiography (TOE) showed the interference of this mass, with a pannus appearance, with the closure of the prosthetic disc. Under conditions of repeated angina episodes, under anticoagulant treatment, surgery was performed, with the intraoperative confirmation of pannus and its removal. Postoperative evolution was favorable. This case reflects the diagnostic and therapeutic management problems of pannus-related prosthetic valve dysfunction.

  7. Towards reducing thrombogenicity of LVAD therapy: optimizing surgical and patient management strategies

    Science.gov (United States)

    Chivukula, Venkat Keshav; Lafzi, Ali; Mokadam, Nahush; Beckman, Jennifer; Mahr, Claudius; Aliseda, Alberto

    2017-11-01

    Unfavourable hemodynamics in heart failure patients implanted with left ventricular assist devices (LVAD), due to non-optimal surgical configurations and patient management, strongly influence thrombogenicity. This is consistent with the increase in devastating thromboembolic complications (specifically thrombosis and stroke) in patients, even as the risk of thrombosis inside the device decreases with modern designs. Inflow cannula and outflow graft surgical configurations have been optimized via patient-specific modeling that computes the thrombogenic potential with a combination of Eulerian (endothelial) wall shear stress and Lagrangian (platelet shear history) tracking. Using this view of hemodynamics, the benefits of intermittent aortic valve opening (promoting washout and reducing stagnant flow in the aortic valve region) have been assessed in managing the patient's residual native cardiac output. The use of this methodology to understand the contribution of the hemodynamics in the flow surrounding the LVAD itself to thrombogenesis show promise in developing holistic patient-specific management strategies to minimize stroke risk and enhance efficacy of LVAD therapy. Funded in part by an AHA postdoctoral fellowship 16POST30520004.

  8. The bicuspid aortic valve and its relation to aortic dilation

    Directory of Open Access Journals (Sweden)

    Shi-Min Yuan

    2010-01-01

    aortic valve disorders at a younger age than the TAV patients and were predominantly male. Aortic dilation was observed in the aortic root, sinotubular junction, and ascending aortic segments in both the BAV and TAV patients undergoing surgical aortic valve replacement, although the BAV patients had a smaller degree of dilation than the TAV patients, and dilation was also significantly age-related in this group. The shorter left main coronary artery that the BAV patients possess may contribute to the progressive course of aortic dilation that these patients experience. Statin therapy did not affect the aortic annulus in either group, but did decrease the dimensions of the aortic root, sinotubular junction and ascending aorta. In general, statin therapy had a better effect on the aortas of the TAV patients than it did on those of the BAV patients.

  9. Pannus-Related Mechanical Valve Dysfunction Leading to Hemodynamic Shock

    Directory of Open Access Journals (Sweden)

    Manabu Shiraishi

    2012-02-01

    Full Text Available Mechanical prosthetic valve dysfunction caused by pannus formation is rare. Pannus restricts movement of prosthetic valve leaflets, resulting in severe aortic regurgitation. We describe the case of a 77-year-old woman who presented to the emergency room with increasing dyspnea, ischemia, and shock secondary to mechanical aortic valve dysfunction. Transesophageal echocardiography showed a blockade of the leaflets of the mechanical aortic valve, with severe aortic regurgitation. She underwent emergent cardiac surgery for aortic valve replacement. Pannus formation should be considered as a potential cause of acute severe aortic regurgitation in a patient with a small-sized mechanical aortic prosthesis in the supra-annular position. On a pathological exam, extensive pannus was found on the ventricular side of the prosthetic valve, extending from the ring into the central orifice. [Arch Clin Exp Surg 2012; 1(1.000: 50-53

  10. Relation of mitral valve morphology and motion to mitral regurgitation severity in patients with mitral valve prolapse

    Directory of Open Access Journals (Sweden)

    Sénéchal Mario

    2012-01-01

    Full Text Available Abstract Background Mitral valve thickness is used as a criterion to distinguish the classical from the non-classical form of mitral valve prolapse (MVP. Classical form of MVP has been associated with higher risk of mitral regurgitation (MR and concomitant complications. We sought to determine the relation of mitral valve morphology and motion to mitral regurgitation severity in patients with MVP. Methods We prospectively analyzed transthoracic echocardiograms of 38 consecutive patients with MVP and various degrees of MR. In the parasternal long-axis view, leaflets length, diastolic leaflet thickness, prolapsing depth, billowing area and non-coaptation distance between both leaflets were measured. Results Twenty patients (53% and 18 patients (47% were identified as having moderate to severe and mild MR respectively (ERO = 45 ± 27 mm2 vs. 5 ± 7 mm2, p Conclusions In patients with MVP, thick mitral leaflet is not associated with significant MR. Leaflet thickness is probably not as important in risk stratification as previously reported in patients with MVP. Other anatomical and geometrical features of the mitral valve apparatus area appear to be much more closely related to MR severity.

  11. The influence of biomaterials on endothelial cell thrombogenicity

    Science.gov (United States)

    McGuigan, Alison P.; Sefton, Michael V.

    2007-01-01

    Driven by tissue engineering and regenerative medicine, endothelial cells are being used in combination with biomaterials in a number of applications for the purpose of improving blood compatibility and host integration. Endothelialized vascular grafts are beginning to be used clinically with some success in some centers, while endothelial seeding is being explored as a means of creating a vasculature within engineered tissues. The underlying assumption of this strategy is that when cultured on artificial biomaterials, a confluent layer of endothelial cells maintain their non-thrombogenic phenotype. In this review the existing knowledge base of endothelial cell thrombogenicity cultured on a number of different biomaterials is summarized. The importance of selecting appropriate endpoint measures that are most reflective of overall surface thrombogenicity is the focus of this review. Endothelial cells inhibit thrombosis through three interconnected regulatory systems (1) the coagulation cascade (2) the cellular components of the blood such as leukocytes and platelets and (3) the complement cascade, and also through effects on fibrinolysis and vascular tone, the latter which influences blood flow. Thus, in order to demonstrate the thromobgenic benefit of seeding a biomaterial with EC, the conditions under which EC surfaces are more likely to exhibit lower thrombogenicity than unseeded biomaterial surfaces need to be consistent with the experimental context. The endpoints selected should be appropriate for the dominant thrombotic process that occurs under the given experimental conditions. PMID:17316788

  12. Check valve

    Science.gov (United States)

    Upton, H.A.; Garcia, P.

    1999-08-24

    A check valve for use in a GDCS of a nuclear reactor and having a motor driven disk including a rotatable armature for rotating the check valve disk over its entire range of motion is described. In one embodiment, the check valve includes a valve body having a coolant flow channel extending therethrough. The coolant flow channel includes an inlet end and an outlet end. A valve body seat is located on an inner surface of the valve body. The check valve further includes a disk assembly, sometimes referred to as the motor driven disc, having a counterweight and a disk shaped valve. The disk valve includes a disk base having a seat for seating with the valve body seat. The disk assembly further includes a first hinge pin member which extends at least partially through the disk assembly and is engaged to the disk. The disk valve is rotatable relative to the first hinge pin member. The check valve also includes a motor having a stator frame with a stator bore therein. An armature is rotatably positioned within the stator bore and the armature is coupled to the disk valve to cause the disk valve to rotate about its full range of motion. 5 figs.

  13. Check valve

    International Nuclear Information System (INIS)

    Upton, H.A.; Garcia, P.

    1999-01-01

    A check valve for use in a GDCS of a nuclear reactor and having a motor driven disk including a rotatable armature for rotating the check valve disk over its entire range of motion is described. In one embodiment, the check valve includes a valve body having a coolant flow channel extending therethrough. The coolant flow channel includes an inlet end and an outlet end. A valve body seat is located on an inner surface of the valve body. The check valve further includes a disk assembly, sometimes referred to as the motor driven disc, having a counterweight and a disk shaped valve. The disk valve includes a disk base having a seat for seating with the valve body seat. The disk assembly further includes a first hinge pin member which extends at least partially through the disk assembly and is engaged to the disk. The disk valve is rotatable relative to the first hinge pin member. The check valve also includes a motor having a stator frame with a stator bore therein. An armature is rotatably positioned within the stator bore and the armature is coupled to the disk valve to cause the disk valve to rotate about its full range of motion. 5 figs

  14. Optimized maintenance concept of safety relevant valves related to ageing management features in nuclear power plants

    International Nuclear Information System (INIS)

    Koring, R.

    2007-01-01

    This paper presents the existing concept in E.ON Kernkraft and its sound application to ageing management issues by focussing on group 2 components such as safety relevant valves. It is demonstrated how the maintenance concept of safety relevant valves is supported by a valve diagnostic system accompanied by an applied procedure to assess the measured results with respect to the required functionality and ageing phenomena. Furthermore this concept has been developed to optimize the existing preventive maintenance of the safety relevant valves by implementing condition oriented aspects derived from the diagnostic results. The main issue of this maintenance concept is to demonstrate the high level of the secured function, reliability and performance of the safety relevant valves within an integrated ageing management. Additionally it offers improvements of all preventive maintenance issues as maintenance periods and the component related volume, spare parts management and costs. (author)

  15. IEEE Std 382-1980: IEEE standard for qualification of safety-related valve actuators

    International Nuclear Information System (INIS)

    Anon.

    1992-01-01

    This standard describes the qualification of all types of power-driven valve actuators, including damper actuators, for safety-related functions in nuclear power generating stations. This standard may also be used to separately qualify actuator components. This standard establishes the minimum requirements for, and guidance regarding, the methods and procedures for qualification of all safety-related functions of power-driven valve actuators

  16. Engineering and maintenance applied to safety-related valves in nuclear power plants

    International Nuclear Information System (INIS)

    Verdu, M. F.; Perez-Aranda, J.

    2014-01-01

    Nuclear Division in Iberdrola engineering and Construction has a team with extensive experience on engineering and services works related to valves. Also, this team is linked to UNESA as Technical support and Reference Center. Iberdrola engineering and construction experience in nuclear power plants valves, gives effective response to engineering and maintenance works that can be demanded in a nuclear power plant and it requires a high degree of qualification and knowledge both in Operation and Outages. (Author)

  17. Tricuspid valve endocarditis following central venous cannulation: The increasing problem of catheter related infection

    Directory of Open Access Journals (Sweden)

    Suresh Babu Kale

    2013-01-01

    Full Text Available A central venous catheter (CVC is inserted for measurement of haemodynamic variables, delivery of nutritional supplements and drugs and access for haemodialysis and haemofiltration. Catheterization and maintenance are common practices and there is more to the technique than routine placement as evident when a procedure-related complication occurs. More than 15% of the patients who receive CVC placement have some complications and infectious endocarditis involving the tricuspid valve is a rare and serious complication with high morbidity and mortality. Overenthusiastic and deep insertion of the guide wire and forceful injection through the CVC may lead to injury of the tricuspid valve and predispose to bacterial deposition and endocarditis. We report a case of tricuspid valve endocarditis, probably secondary to injury of the anterior tricuspid leaflet by the guide wire or the CVC that required open heart surgery with vegetectomy and repair of the tricuspid valve.

  18. Stent fracture, valve dysfunction, and right ventricular outflow tract reintervention after transcatheter pulmonary valve implantation: patient-related and procedural risk factors in the US Melody Valve Trial.

    Science.gov (United States)

    McElhinney, Doff B; Cheatham, John P; Jones, Thomas K; Lock, James E; Vincent, Julie A; Zahn, Evan M; Hellenbrand, William E

    2011-12-01

    Among patients undergoing transcatheter pulmonary valve (TPV) replacement with the Melody valve, risk factors for Melody stent fracture (MSF) and right ventricular outflow tract (RVOT) reintervention have not been well defined. From January 2007 to January 2010, 150 patients (median age, 19 years) underwent TPV implantation in the Melody valve Investigational Device Exemption trial. Existing conduit stents from a prior catheterization were present in 37 patients (25%, fractured in 12); 1 or more new prestents were placed at the TPV implant catheterization in 51 patients. During follow-up (median, 30 months), MSF was diagnosed in 39 patients. Freedom from a diagnosis of MSF was 77±4% at 14 months (after the 1-year evaluation window) and 60±9% at 39 months (3-year window). On multivariable analysis, implant within an existing stent, new prestent, or bioprosthetic valve (combined variable) was associated with longer freedom from MSF (Pbioprosthetic valve was associated with lower risk of MSF and reintervention.

  19. Survival and freedom from aortic valve-related reoperation after valve-sparing aortic root replacement in 1015 patients.

    Science.gov (United States)

    Kari, Fabian A; Doll, Kai-Nicolas; Hemmer, Wolfgang; Liebrich, Markus; Sievers, Hans-Hinrich; Richardt, Doreen; Reichenspurner, Hermann; Detter, Christian; Siepe, Matthias; Czerny, Martin; Beyersdorf, Friedhelm

    2016-04-01

    The aim of this study was to characterize mortality and aortic valve replacement after valve-sparing aortic root replacement (V-SARR) in a multicentre cohort. Between 1994 and 2014, 1015 patients had V-SARR with (n = 288, 28%) or without cusp/commissure repair (n = 727, 72%) at the centres of Lübeck (n = 343, 34%), Stuttgart (n = 346, 34%), Hamburg (n = 109, 11%) and Freiburg (n = 217, 21%), Germany. Comparative survival of an age- and gender-matched general population was calculated. Log-rank tests and multiple logistic regression were used to identify risk factors. The mean follow-up was 5.2 ± 3.9 years. Cumulative follow-up comprised 2933 patient-years. Early survival was 98%. NYHA status and aneurysm size were predictive of death during mid-term follow-up (P = 0.025). Freedom from aortic valve replacement was 90% at 8 years, with the type of V-SARR (root remodelling, David II) being a risk factor (P = 0.015). Bicuspid aortic valve (P = 0.26) and initial valve function (P = 0.4) did not impact reoperation. The need of additional valve repair (cusps/commissures) was not linked to reoperation: freedom from aortic valve replacement at 8 years was 84% if cusp repair was performed versus 90% if V-SARR alone was performed (P = 0.218). Marfan syndrome had no impact on survival or on aortic valve replacement. Mid-term survival of patients after V-SARR is comparable with that of a matched general population. The regurgitant bicuspid aortic valve is a favourable substrate for V-SARR. Prophylactic surgery should be performed before symptoms or large aneurysms are present to achieve optimal mid-term outcomes. © The Author 2015. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.

  20. Thrombogenicity tests on ar-irradiated polycarbonate foils

    Energy Technology Data Exchange (ETDEWEB)

    Trindade, Gustavo F.; Rizzutto, Marcia A.; Silva, Tiago F.; Moro, Marcos V.; Added, Nemitala; Tabacniks, Manfredo H., E-mail: g.ferraz@usp.br [Universidade de Sao Paulo (USP), Sao Paulo, SP (Brazil). Inst. de Fisica; Delgado, Adriana O. [Universidade Federal de Sao Carlos (UFSCAR), Sorocaba, SP (Brazil); Cunha, Tatiana F. [Biosintesis P and D do Brasil, Sao Paulo, SP (Brazil); Higa, Olga Z. [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil). Centro de Biotecnologia

    2013-07-01

    Understanding polymer surface properties is extremely important for the most wide range of their applications, from basic coating to the most complex composites and biomaterials. Low energy ion beam irradiation of polymer can improve such surface properties. By modifying its surface biocompatibility, polymers are excellent candidates for biomaterials, due to its malleability and low weight, when compared to metals. In this work, we irradiated 30-μm Bisphenol-A Polycarbonate foils with 23-keV Argon ion beam at six different doses. Aluminium foils were simultaneously irradiated in order to measure the doses by Rutherford Backscattering Spectroscopy. The surface modifications after the argon ion beam irradiation were analyzed by water contact angle measurements and atomic force microscopy. Platelet adhesion tests were used in order to investigate thrombogenicity, showing a growing tendency with the irradiated Argon dose. (author)

  1. Thrombogenicity tests on ar-irradiated polycarbonate foils

    International Nuclear Information System (INIS)

    Trindade, Gustavo F.; Rizzutto, Marcia A.; Silva, Tiago F.; Moro, Marcos V.; Added, Nemitala; Tabacniks, Manfredo H.; Cunha, Tatiana F.; Higa, Olga Z.

    2013-01-01

    Understanding polymer surface properties is extremely important for the most wide range of their applications, from basic coating to the most complex composites and biomaterials. Low energy ion beam irradiation of polymer can improve such surface properties. By modifying its surface biocompatibility, polymers are excellent candidates for biomaterials, due to its malleability and low weight, when compared to metals. In this work, we irradiated 30-μm Bisphenol-A Polycarbonate foils with 23-keV Argon ion beam at six different doses. Aluminium foils were simultaneously irradiated in order to measure the doses by Rutherford Backscattering Spectroscopy. The surface modifications after the argon ion beam irradiation were analyzed by water contact angle measurements and atomic force microscopy. Platelet adhesion tests were used in order to investigate thrombogenicity, showing a growing tendency with the irradiated Argon dose. (author)

  2. In vitro 2D PIV measurements and related aperture areas of tricuspid bioprosthetic mitral valves at the beginning of diastole.

    Science.gov (United States)

    Bazan, Ovandir; Ortiz, Jayme Pinto; Fukumasu, Newton Kiyoshi; Pacifico, Antonio Luiz; Yanagihara, Jurandir Itizo

    2016-07-04

    Besides ventricular parameters, the design and angular orientation of a prosthetic heart valve induce a specific flow field. The aim of this study was to know the inflow characteristics of a left ventricular model (LVM), investigating the behavior of tricuspid bioprosthetic mitral valves in terms of velocity profiles and related valve aperture areas at the beginning of diastole, under different conditions. 3 heart rates (HRs) were established in the LVM and each mitral bioprosthesis (27 and 31 mm diameter) was installed in 2 orientations, rotated by 180° . For each experimental setup, 2-dimensional particle image velocimetry (2D PIV) measurements and simultaneous mitral valve (MV) area detection were obtained from 50 samples. The results from the velocity profiles immediately downstream of mitral bioprostheses showed the influence of valve orientation for moderate HRs, although for a similar magnitude of mean velocity vectors. The geometries of MV open areas for each HR were similar regardless of valve orientation, except for the 27-mm valve at 90 beats per minute (bpm), and for the 31-mm valve at 60 bpm. Moreover, for each HR, similar percentages of valve open area were obtained regardless of MV nominal diameters. In conclusion, the experimental setup for the 2D PIV measurements synchronized with the MV area detection was a useful tool for knowing the inflow characteristics of the LVM.

  3. Altered aortic shape in bicuspid aortic valve relatives influences blood flow patterns.

    Science.gov (United States)

    Schnell, Susanne; Smith, Danielle A; Barker, Alex J; Entezari, Pegah; Honarmand, Amir R; Carr, Maria L; Malaisrie, S Chris; McCarthy, Patrick M; Collins, Jeremy; Carr, James C; Markl, Michael

    2016-11-01

    Bicuspid aortic valve (BAV) is known to exhibit familial inheritance and is associated with aortopathy and altered aortic haemodynamics. However, it remains unclear whether BAV-related aortopathy can be inherited independently of valve morphology. Four-dimensional flow magnetic resonance imaging for the in vivo assessment of thoracic aortic 3D blood flow was performed in 24 BAV relatives with trileaflet aortic valves (age = 40 ± 14 years) and 15 healthy controls (age = 37 ± 10 years). Data analysis included aortic dimensions, shape (round/gothic/cubic), and 3D blood flow characteristics (semi-quantitative vortex/helix grading and peak velocities). Cubic and gothic aortic shapes were markedly more prevalent in BAV relatives compared with controls (38 vs. 7%). Ascending aorta (AAo) vortex flow in BAV relatives was significantly increased compared with controls (grading = 1.5 ± 1.0 vs. 0.6 ± 0.9, P = 0.015). Aortic haemodynamics were influenced by aortic shape: peak velocities were reduced for gothic aortas vs. round aortas (P = 0.003); vortex flow was increased for cubic aortas in the AAo (P gothic aortas in the AAo and descending aorta (P = 0.003, P = 0.029). Logistic regression demonstrated significant associations of shape with severity of vortex flow in AAo (P < 0.001) and aortic arch (P = 0.016) in BAV relatives. BAV relatives expressed altered aortic shape and increased vortex flow despite the absence of valvular disease or aortic dilatation. These data suggest a heritable component of BAV-related aortopathy affecting aortic shape and aberrant blood flow, independent of valve morphology. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.

  4. New insights into valve-related intramural and intracellular bacterial diversity in infective endocarditis.

    Science.gov (United States)

    Oberbach, Andreas; Schlichting, Nadine; Feder, Stefan; Lehmann, Stefanie; Kullnick, Yvonne; Buschmann, Tilo; Blumert, Conny; Horn, Friedemann; Neuhaus, Jochen; Neujahr, Ralph; Bagaev, Erik; Hagl, Christian; Pichlmaier, Maximilian; Rodloff, Arne Christian; Gräber, Sandra; Kirsch, Katharina; Sandri, Marcus; Kumbhari, Vivek; Behzadi, Armirhossein; Behzadi, Amirali; Correia, Joao Carlos; Mohr, Friedrich Wilhelm; Friedrich, Maik

    2017-01-01

    In infective endocarditis (IE), a severe inflammatory disease of the endocardium with an unchanged incidence and mortality rate over the past decades, only 1% of the cases have been described as polymicrobial infections based on microbiological approaches. The aim of this study was to identify potential biodiversity of bacterial species from infected native and prosthetic valves. Furthermore, we compared the ultrastructural micro-environments to detect the localization and distribution patterns of pathogens in IE. Using next-generation sequencing (NGS) of 16S rDNA, which allows analysis of the entire bacterial community within a single sample, we investigated the biodiversity of infectious bacterial species from resected native and prosthetic valves in a clinical cohort of 8 IE patients. Furthermore, we investigated the ultrastructural infected valve micro-environment by focused ion beam scanning electron microscopy (FIB-SEM). Biodiversity was detected in 7 of 8 resected heart valves. This comprised 13 bacterial genera and 16 species. In addition to 11 pathogens already described as being IE related, 5 bacterial species were identified as having a novel association. In contrast, valve and blood culture-based diagnosis revealed only 4 species from 3 bacterial genera and did not show any relevant antibiotic resistance. The antibiotics chosen on this basis for treatment, however, did not cover the bacterial spectra identified by our amplicon sequencing analysis in 4 of 8 cases. In addition to intramural distribution patterns of infective bacteria, intracellular localization with evidence of bacterial immune escape mechanisms was identified. The high frequency of polymicrobial infections, pathogen diversity, and intracellular persistence of common IE-causing bacteria may provide clues to help explain the persistent and devastating mortality rate observed for IE. Improved bacterial diagnosis by 16S rDNA NGS that increases the ability to tailor antibiotic therapy may

  5. A discussion of system reliability and the relative importance of pumps and valves to overall system availability

    Energy Technology Data Exchange (ETDEWEB)

    Poole, A.B. [Oak Ridge National Lab., TN (United States)

    1996-12-01

    An analysis was undertaken to establish preliminary trends for how component aging can effect failure rates for swing check valves, centrifugal pumps and motor operated valves. These failure rate trends were evaluated over time and linear aging rate models established. The failure rate models were then used with classic reliability theories to estimate reliability as a function of operating time. Reliability theory was also used to establish a simple system reliability model. Using the system model, the relative importance of pumps and valves to the overall system reliability were studied. Conclusions were established relative to overall system availability over time and the relative unavailabilities of the various components studied.

  6. A discussion of system reliability and the relative importance of pumps and valves to overall system availability

    International Nuclear Information System (INIS)

    Poole, A.B.

    1996-01-01

    An analysis was undertaken to establish preliminary trends for how component aging can effect failure rates for swing check valves, centrifugal pumps and motor operated valves. These failure rate trends were evaluated over time and linear aging rate models established. The failure rate models were then used with classic reliability theories to estimate reliability as a function of operating time. Reliability theory was also used to establish a simple system reliability model. Using the system model, the relative importance of pumps and valves to the overall system reliability were studied. Conclusions were established relative to overall system availability over time and the relative unavailabilities of the various components studied

  7. Quantitation of thrombogenicity of hemodialyzer with technetium-99m and indium-111 labeled platelets

    International Nuclear Information System (INIS)

    Dewanjee, M.K.; Kapadvanjwala, M.; Ruzius, K.; Serafini, A.N.; Zilleruelo, G.E.; Sfakianakis, G.N.

    1993-01-01

    The platelet thrombogenicity of a hemodialyzer was quantified with 99m Tc- and 111 In-labeled platelets. The platelets collected from blood of Beagle dogs, Yorkshire pigs and humans were labeled with 111 In-tropolone (detergent-free) and 99m Tc-HMPAO. Hemodialysis was performed with a hollow-fiber dialyzer (HFD) in a flow-loop, the temperature maintained at 37 o C, with flow-rates of 7, 150 and 270 mL/min; after dialysis, HFD radioactivity was measured with an ionization chamber and imaged with a γ-camera. The dynamic processes of thrombosis and embolization from the hemodialyzer resulted in large standard deviations around the mean values of the adherent thrombus. In spite of this, the quantitation of comparative thrombogenicity with 111 In-and 99m Tc-labeled platelets suggests that both radionuclides could be used for measurement of device-induced thrombogenicity and estimation of prosthesis-induced thrombogenicity. (author)

  8. Relation of Bicuspid Aortic Valve Morphology to the Dilatation Pattern of the Proximal Aorta: Focus on the Transvalvular Flow

    Directory of Open Access Journals (Sweden)

    Evaldas Girdauskas

    2012-01-01

    Full Text Available Whether the dilatation of proximal aorta in patients with bicuspid aortic valve is secondary to hemodynamic effects related to the abnormal aortic valve or a primary manifestation of the genetic disorder remains controversial. We discuss in this paper the recent data on the BAV function and transvalvular flow patterns in relation with the dilatation type of the proximal aorta. Different morphological forms of bicuspid aortic valve in relation with the specific transvalvular blood flow patterns are focus of the first paragraph of this paper. In the second part of this paper we present the pathogenetic insight into the different clinically observed phenotypes of bicuspid aortic valve disease (i.e., association of proximal aortic shapes with the specific cusp fusion patterns, based on the data from recent rheological studies.

  9. Corrosion of valve metals

    International Nuclear Information System (INIS)

    Draley, J.E.

    1976-01-01

    A general survey related to the corrosion of valve metals or film-forming metals. The way these metals corrode with some general examples is described. Valve metals form relatively perfect oxide films with little breakdown or leakage when anodized

  10. Regulatory standpoints on the design-basis capability of safety-related motor-operated valves(MOVs) and power-operated gate valves(POGVs)

    International Nuclear Information System (INIS)

    Kim, W. T.; Kum, O. H.

    1999-01-01

    The weakness in the design-basis capability of Motor-Operated Valves(MOVs) and the susceptibility to Pressure Locking and Thermal Binding phenomena of Power-Operated Gate Valves(POGVs) have been major concerns to be resolved in the nuclear society in and abroad since Three Mile Island accident occurred in the USA in 1979. Through detailed analysis of operating experience and regulatory activities, some MOVs and POGVs have been found to be unreliable in performing their safety functions when they are required to do so under certain conditions, especially under design-basis accident conditions. Further, it is well understood that these safety problems may not be identified by the typical valve in-service testing(IST). USNRC has published three Generic Letters, GL 89-10, GL 95-07, and GL 96-05, requiring nuclear plant licensees to take appropriate actions to resolve the problems mentioned above. Korean nuclear regulatory body has made public an administration measure called 'Regulatory recommendation to verify safety functions of the safety-related MOVs and POGVs' on June 13, 1997, and in this administration measure Korean utility is asked to submit written documents to show how it assure design-basis capability of these valves. The following are among the major concerns being considered from a regulation standpoint. Program scope and implementation priority, dynamic tests under differential pressure conditions, accuracy of diagnostic equipment, torque switch setting and torque bypass percentage, weak link analysis, motor actuator sizing, corrective actions taken to resolve pressure locking and thermal binding susceptibility, and a periodic verification program for the valves once design-basis capability has been verified

  11. Relation of thoracic aortic and aortic valve calcium to coronary artery calcium and risk assessment.

    Science.gov (United States)

    Wong, Nathan D; Sciammarella, Maria; Arad, Yadon; Miranda-Peats, Romalisa; Polk, Donna; Hachamovich, Rory; Friedman, John; Hayes, Sean; Daniell, Anthony; Berman, Daniel S

    2003-10-15

    Aortic calcium, aortic valve calcium (AVC), and coronary artery calcium (CAC) have been associated with cardiovascular event risk. We examined the prevalence of thoracic aortic calcium (TAC) and AVC in relation to the presence and extent of CAC, cardiovascular risk factors, and estimated risk of coronary heart disease (CHD). In 2,740 persons without known CHD aged 20 to 79 years, CAC was assessed by electron beam- or multidetector-computed tomography. We determined the prevalence of TAC and AVC in relation to CAC, CHD risk factors, and predicted 10-year risk of CHD. A close correspondence of TAC and AVC was observed with CAC. TAC and AVC increased with age; by the eighth decade of life, the prevalence of TAC was similar to that of CAC (>80%), and 36% of men and 24% of women had AVC. Age, male gender, and low-density lipoprotein cholesterol were directly related to the likelihood of CAC, TAC, and AVC; higher diastolic blood pressure and cigarette smoking additionally predicted CAC. Body mass index and higher systolic and lower diastolic blood pressures were also related to TAC, and higher body mass index and lower diastolic blood pressure were related to AVC. Calculated risk of CHD increased with the presence of AVC and TAC across levels of CAC. TAC and AVC provided incremental value over CAC in association with the 10-year calculated risk of CHD. If longitudinal studies show an incremental value of aortic and aortic valve calcium over that of CAC for prediction of cardiovascular events, future guidelines for risk assessment incorporating CAC assessment may additionally incorporate the measurement of aortic and/or aortic valve calcium.

  12. Diseases of the Tricuspid Valve

    Science.gov (United States)

    ... stenosis. Tricuspid Regurgitation Tricuspid regurgitation is also called tricuspid insufficiency or tricuspid incompetence. It means there is a ... require valve surgery. Tags: heart valves , tricuspid incompetence , ... tricuspid regurgitation , tricuspid stenosis , valve disease Related Links ...

  13. Survey of valve operator-related events occurring during 1978, 1979 and 1980

    International Nuclear Information System (INIS)

    Brown, E.J.; Ashe, F.S.

    1983-01-01

    The survey approach was to analyze several events and identify trends or patterns. The primary data source was licensee event reports (LERs) and consisted of 444 total valve operator events with 193 motor operator events which served as the basis for this study. The investigation revealed that motor-operated events could be grouped in three major categories which are torque switches, limit switches, and motors. The major findings are: (1) Torque switches do not appear to be a dominant cause of valve assembly inoperability. The reported information suggests torque switch events are an indication of symptomatic change with time in valve operability characteristics rather than a root cause of valve inoperability. (2) Repetitive problems are occurring with valve operators. It may occur on the same valve, a valve in similar service in a similar system, or a valve in similar service in a redundant train of the same system. (3) The plant operating staff objective appears to be a mode of finding measures to return inoperable equipment to operational status rather than to determine root causes of inoperability. (4) Motor burnout of valve motor operators has occurred quite frequently in High Pressure Coolant Injection (HPCI) and Reactor Core Isolation Cooling (RCIC) systems of BWR units. (orig./GL)

  14. Categorization of safety related motor operated valve safety significance for Ulchin Unit 3

    International Nuclear Information System (INIS)

    Kang, D. I.; Kim, K. Y.

    2002-03-01

    We performed a categorization of safety related Motor Operated Valve (MOV) safety significance for Ulchin Unit 3. The safety evaluation of MOV of domestic nuclear power plants affects the generic data used for the quantification of MOV common cause failure ( CCF) events in Ulchin Units 3 PSA. Therefore, in this study, we re-estimated the MGL(Multiple Greek Letter) parameter used for the evaluation of MOV CCF probabilities in Ulchin Units 3 Probabilistic Safety Assessment (PSA) and performed a classification of the MOV safety significance. The re-estimation results of the MGL parameter show that its value is decreased by 30% compared with the current value in Ulchin Unit 3 PSA. The categorization results of MOV safety significance using the changed value of MGL parameter shows that the number of HSSCs(High Safety Significant Components) is decreased by 54.5% compared with those using the current value of it in Ulchin Units 3 PSA

  15. REDUCED THROMBOGENICITY OF VASCULAR PROSTHESES BY COATING WITH ADP-ASE

    NARCIS (Netherlands)

    VANDERLEI, B; ROBINSON, PH; BAKKER, WW; Bartels, H.

    1992-01-01

    In this pilot study ADP-ase coated polyurethane (PL) vascular prostheses and noncoated (control) PU vascular prostheses (all vascular prostheses: ID 1.5 mm, length 1,5 cm) were implanted into the carotid artery of the rabbit to test wheter ADP-ase might function as an adequate anti-thrombogenic

  16. Patients' self-reported function, symptoms and health-related quality of life before and 6 months after transcatheter aortic valve implantation and surgical aortic valve replacement.

    Science.gov (United States)

    Olsson, Karin; Nilsson, Johan; Hörnsten, Åsa; Näslund, Ulf

    2017-03-01

    Aortic stenosis is the most common valve disease in western countries and has poor prognosis without treatment. Surgical aortic valve replacement (SAVR) is the gold standard, and transcatheter aortic valve implantation (TAVI) is a new method that is used in high-risk patients who are denied surgery. The purpose of treatment is not only to save life, but also to reduce symptoms and increase health-related quality of life (HRQoL). The aim of this study was to describe patients' self-reported outcomes in terms of physical function, symptoms, dependence, HRQoL, and cognitive function after TAVI and SAVR. All patients treated with TAVI during 1 year ( n = 24) and age-matched patients treated with SAVR ( n = 24) were included. Data were collected on the day before and at 6 months after treatment using structural questionnaires. Self-rated function was low before treatment and increased at follow-up. A quarter of all patients reported syncope at baseline, and none reported this at follow-up. Breathlessness was reported by all patients to be the most limiting cardiac symptom, but the TAVI patients reported more severe symptoms. At 6 months' follow-up, symptoms were reduced, but breathlessness and fatigue were still common, especially in the TAVI group. HRQoL, which was very low in the TAVI group at baseline, increased in all dimensions except social function. We found no change in cognitive function or dependence at follow-up. There was no difference in the size of improvement between groups. The results could be helpful when informing future patients in order to give them realistic expectations.

  17. Heart valve surgery

    Science.gov (United States)

    ... replacement; Valve repair; Heart valve prosthesis; Mechanical valves; Prosthetic valves ... surgery. Your heart valve has been damaged by infection ( endocarditis ). You have received a new heart valve ...

  18. Valve Disease

    Science.gov (United States)

    ... blood. There are 4 valves in the heart: tricuspid, pulmonary, mitral, and aortic. Two types of problems can disrupt blood flow through the valves: regurgitation or stenosis. Regurgitation is also called insufficiency or incompetence. Regurgitation happens when a valve doesn’ ...

  19. Prediction of thrombus-related mechanical prosthetic valve dysfunction using transesophageal echocardiography

    Science.gov (United States)

    Lin, S. S.; Tiong, I. Y.; Asher, C. R.; Murphy, M. T.; Thomas, J. D.; Griffin, B. P.

    2000-01-01

    Identification of thrombus-related mechanical prosthetic valve dysfunction (MPVD) has important therapeutic implications. We sought to develop an algorithm, combining clinical and echocardiographic parameters, for prediction of thrombus-related MPVD in a series of 53 patients (24 men, age 52 +/- 16 years) who had intraoperative diagnosis of thrombus or pannus from 1992 to 1997. Clinical and echocardiographic parameters were analyzed to identify predictors of thrombus and pannus. Prevalence of thrombus and diagnostic yields relative to the number of predictors were determined. There were 22 patients with thrombus, 19 patients with pannus, and 12 patients with both. Forty-two of 53 masses were visualized using transesophageal echocardiography (TEE), including 29 of 34 thrombi or both thrombi and panni and 13 of 19 isolated panni. Predictors of thrombus or mixed presentation include mobile mass (p = 0.009), attachment to occluder (p = 0.02), elevated gradients (p = 0.04), and an international normalized ratio of or = 1 predictor. The prevalence of thrombus in the presence of or = 3 predictors is 14%, 69%, and 91%, respectively. Thus, TEE is sensitive in the identification of abnormal mass in the setting of MPVD. An algorithm based on clinical and transesophageal echocardiographic predictors may be useful to estimate the likelihood of thrombus in the setting of MPVD. In the presence of > or = 3 predictors, the probability of thrombus is high.

  20. Transcatheter Aortic Valve Implantation and Morbidity and Mortality-Related Factors: a 5-Year Experience in Brazil

    Directory of Open Access Journals (Sweden)

    André Luiz Silveira Souza

    2016-01-01

    Full Text Available Abstract Background: Transcatheter aortic valve implantation has become an option for high-surgical-risk patients with aortic valve disease. Objective: To evaluate the in-hospital and one-year follow-up outcomes of transcatheter aortic valve implantation. Methods: Prospective cohort study of transcatheter aortic valve implantation cases from July 2009 to February 2015. Analysis of clinical and procedural variables, correlating them with in-hospital and one-year mortality. Results: A total of 136 patients with a mean age of 83 years (80-87 underwent heart valve implantation; of these, 49% were women, 131 (96.3% had aortic stenosis, one (0.7% had aortic regurgitation and four (2.9% had prosthetic valve dysfunction. NYHA functional class was III or IV in 129 cases (94.8%. The baseline orifice area was 0.67 ± 0.17 cm2 and the mean left ventricular-aortic pressure gradient was 47.3±18.2 mmHg, with an STS score of 9.3% (4.8%-22.3%. The prostheses implanted were self-expanding in 97% of cases. Perioperative mortality was 1.5%; 30-day mortality, 5.9%; in-hospital mortality, 8.1%; and one-year mortality, 15.5%. Blood transfusion (relative risk of 54; p = 0.0003 and pulmonary arterial hypertension (relative risk of 5.3; p = 0.036 were predictive of in-hospital mortality. Peak C-reactive protein (relative risk of 1.8; p = 0.013 and blood transfusion (relative risk of 8.3; p = 0.0009 were predictive of 1-year mortality. At 30 days, 97% of patients were in NYHA functional class I/II; at one year, this figure reached 96%. Conclusion: Transcatheter aortic valve implantation was performed with a high success rate and low mortality. Blood transfusion was associated with higher in-hospital and one-year mortality. Peak C-reactive protein was associated with one-year mortality.

  1. Tight valve

    International Nuclear Information System (INIS)

    Guedj, F.

    1987-01-01

    This sealed valve is made with a valve seat, an axial valve with a rod fixed to its upper end, a thick bell surrounding the rod and welded by a thin join on the valve casing, a threated ring screwed onto the upper end of the rod and a magnet or electromagnet rotating the ring outside the bell [fr

  2. Quantitation of thrombogenicity of hemodialyzer with technetium-99m and indium-111 labeled platelets

    International Nuclear Information System (INIS)

    Dewanjee, M.K.; Kapadvanjwala, Mansoor; Ruzius, Kees; Serafini, A.N.; Zilleruelo, G.E.; Sfakianakis, G.N.

    1993-01-01

    The platelet thromobogenicity of a hemodialyzer was quantified with 99m Tc- and 111 In-labeled platelets. The platelets collected from blood of Beagle dogs, Yorkshire pigs and human volunteers were labeled with 111 in-tropolone (detergent-free) and 99m Tc-HMPAO. Hemodialysis was performed with a hollow-fiber dialyzer (HFD) in a flow-loop, the temperature of which was maintained at 37 o C, with flow-rates of 7, 150 and 270 mL/min; after dialysis, the HFD radioactivity was measured with an ionization chamber and imaged with a γ-camera. The radioactivity of samples of hollow-fibers taken from the top, middle and bottom of the dialyzer was determined with a γ-counter. The mean values of hemodialyzer-adherent platelet radioactivity were calculated for both radionuclides. The canine platelets were found to be more thrombogenic than porcine and human platelets. The adhesivity of porcine platelets to the biomaterial (cellulose-acetate) of the dialyzer approximated that of human platelets. The 99m Tc label underestimated the thrombus formation (P 111 In- and 99m Tc-labeled platelets suggests that both radionuclides could be used for measurement of device-induced thrombogenicity and may provide an estimation of prosthesis-induced thrombogenicity of human platelets from animal studies. (Author)

  3. Aging and service wear of air-operated valves used in safety-related systems at nuclear power plants

    International Nuclear Information System (INIS)

    Cox, D.F.; McElhaney, K.L.; Staunton, R.H.

    1995-05-01

    Air-operated valves (AOVs) are used in a variety of safety-related applications at nuclear power plants. They are often used where rapid stroke times are required or precise control of the valve obturator is required. They can be designed to operate automatically upon loss of power, which is often desirable when selecting components for response to design basis conditions. The purpose of this report is to examine the reported failures of AOVs and determine whether there are identifiable trends in the failures related to predictable causes. This report examines the specific components that comprise a typical AOV, how those components fail, when they fail, and how such failures are discovered. It also examines whether current testing frequencies and methods are effective in predicting such failures

  4. Nuclear valves latest development

    International Nuclear Information System (INIS)

    Isaac, F.; Monier, M.

    1993-01-01

    In the frame of Nuclear Power Plant upgrade (Emergency Power Supply and Emergency Core Cooling), Westinghouse had to face a new valve design philosophy specially for motor operated valves. The valves have to been designed to resist any operating conditions, postulated accident or loss of control. The requirements for motor operated valves are listed and the selected model and related upgrading explained. As part of plant upgrade and valves replacement, Westinghouse has sponsored alternative hardfacing research programme. Two types of materials have been investigated: nickel base alloys and iron base alloys. Programme requirements and test results are given. A new globe valve model (On-Off or regulating) is described developed by Alsthom Velan permitting the seat replacement in less than 10 min. (Z.S.) 2 figs

  5. Long-term tricuspid valve prosthesis-related complications in patients with congenital heart disease

    NARCIS (Netherlands)

    van Slooten, Ymkje J.; Freling, Hendrik G.; van Melle, Joost P.; Mulder, Barbara J. M.; Jongbloed, Monique R. M.; Ebels, Tjark; Voors, Adriaan A.; Pieper, Petronella G.

    2014-01-01

    In patients with acquired valvar disease, morbidity and mortality rates after tricuspid valve replacement (TVR) are high. However, in adult patients with congenital heart disease, though data concerning outcome after TVR are scarce, even poorer results are suggested in patients with Ebstein anomaly.

  6. Long-term tricuspid valve prosthesis-related complications in patients with congenital heart disease

    NARCIS (Netherlands)

    van Slooten, Ymkje J.; Freling, Hendrik G.; van Melle, Joost P.; Mulder, Barbara J. M.; Jongbloed, Monique R. M.; Ebels, Tjark; Voors, Adriaan A.; Pieper, Petronella G.

    OBJECTIVES: In patients with acquired valvar disease, morbidity and mortality rates after tricuspid valve replacement (TVR) are high. However, in adult patients with congenital heart disease, though data concerning outcome after TVR are scarce, even poorer results are suggested in patients with

  7. Aortic valve replacement and the stentless Freedom SOLO valve

    NARCIS (Netherlands)

    Wollersheim, L.W.L.M.

    2016-01-01

    Aortic valve stenosis has become the most prevalent valvular heart disease in Europe and North America, and is generally caused by age-related calcification of the aortic valve. For most patients, severe symptomatic aortic stenosis needs effective mechanical relief in the form of valve replacement

  8. Control Valve

    Energy Technology Data Exchange (ETDEWEB)

    Moore, Wayne R.

    2018-03-20

    A control valve includes a first conduit having a first inlet and a first outlet and defining a first passage; a second conduit having a second inlet and a second outlet and defining a second passage, the second conduit extending into the first passage such that the second inlet is located within the first passage; and a valve plate disposed pivotably within the first passage, the valve plate defining a valve plate surface. Pivoting of the valve plate within the first passage varies flow from the first inlet to the first outlet and the valve plate is pivotal between a first position and a second position such that in the first position the valve plate substantially prevents fluid communication between the first passage and the second passage and such that in the second position the valve plate permits fluid communication between the first passage and the second passage.

  9. Impact of recipient-related factors on structural dysfunction of xenoaortic bioprosthetic heart valves

    Directory of Open Access Journals (Sweden)

    Barbarash O

    2015-03-01

    Full Text Available Olga Barbarash, Natalya Rutkovskaya, Oksana Hryachkova, Olga Gruzdeva, Evgenya Uchasova, Anastasia Ponasenko, Natalya Kondyukova, Yuri Odarenko, Leonid Barbarash Federal State Budgetary Scientific Institution Research Institute for Complex Issues of Cardiovascular Diseases, Kemerovo, Russia Objective: To analyze the influence of recipient-related metabolic factors on the rate of structural dysfunction caused by the calcification of xenoaortic bioprostheses. Materials and methods: We retrospectively analyzed clinical status, calcium–phosphorus metabolism, and nonspecific markers of inflammatory response in bioprosthetic mitral valve recipients with calcific degeneration confirmed by histological and electron microscopic studies (group 1, n=22, and in those without degeneration (group 2, n=48. Results: Patients with confirmed calcification of bioprostheses were more likely to have a severe clinical state (functional class IV in 36% in group 1 versus 15% in group 2, P=0.03 and a longer cardiopulmonary bypass period (112.8±18.8 minutes in group 1 versus 97.2±23.6 minutes in group 2, P=0.02 during primary surgery. Patients in group 1 demonstrated moderate hypovitaminosis D (median 34.0, interquartile range [21.0; 49.4] vs 40 [27.2; 54.0] pmol/L, P>0.05, osteoprotegerin deficiency (82.5 [44.2; 115.4] vs 113.5 [65.7; 191.3] pg/mL, P>0.05 and osteopontin deficiency (4.5 [3.3; 7.7] vs 5.2 [4.1; 7.2] ng/mL, P>0.05, and significantly reduced bone-specific alkaline phosphatase isoenzyme (17.1 [12.2; 21.4] vs 22.3 [15.5; 30.5] U/L, P=0.01 and interleukin-8 levels (9.74 [9.19; 10.09] pg/mL vs 13.17 [9.72; 23.1] pg/mL, P=0.045 compared with group 2, with an overall increase in serum levels of proinflammatory markers. Conclusion: Possible predictors of the rate of calcific degeneration of bioprostheses include the degree of decompensated heart failure, the duration and invasiveness of surgery, and the characteristics of calcium–phosphorus homeostasis in

  10. Valve assembly

    International Nuclear Information System (INIS)

    Sandling, M.

    1981-01-01

    An improved valve assembly, used for controlling the flow of radioactive slurry, is described. Radioactive contamination of the air during removal or replacement of the valve is prevented by sucking air from the atmosphere through a portion of the structure above the valve housing. (U.K.)

  11. Comparison of Acute Thrombogenicity for Magnesium versus Stainless Steel Stents in a Porcine Arteriovenous Shunt Model.

    Science.gov (United States)

    Lipinski, Michael J; Acampado, Eduardo; Cheng, Qi; Adams, Lila; Torii, Sho; Gai, Jiaxiang; Torguson, Rebecca; Hellinga, David G; Joner, Michael; Harder, Claus; Zumstein, Philine; Finn, Aloke V; Kolodgie, Frank D; Virmani, Renu; Waksman, Ron

    2018-05-08

    Aims The aetiology for reduced thrombogenicity of the Magmaris resorbable magnesium scaffold (RMS) when compared with the Absorb bioresorbable vascular scaffold remains unclear. We therefore investigated whether the Magmaris RMS has platelet-repelling properties by comparing its acute thrombogenicity with an equivalent stainless steel stent in an arteriovenous shunt model. Methods and Results An ex vivo porcine carotid jugular arteriovenous shunt was established and connected to Sylgard tubing containing the Magmaris RMS with sirolimus-eluting PLLA coating and an equivalent 316L stainless steel stent with sirolimus-eluting PLLA coating. Six shunts (2 shunt runs per pig) were run comparing the 2 scaffolds (n=9) in alternating order. Nested generalised linear mixed models were employed to compare variables between scaffold groups. Confocal fluorescent microscopy costaining CD61/CD42b demonstrated that the 316L equivalent stent had significantly greater platelet coverage of the total scaffold compared with Magmaris (5.8% vs. 2.8%, Rate ratio 2.21 [1.41, 3.47], p=0.012). Scanning electron microscopy demonstrated significantly greater thrombus deposition on the 316L equivalent stent as a percentage of the total scaffold compared with Magmaris (8.0% vs. 5.3%, p=0.009). Magmaris also had significantly less CD14 positive monocyte deposition and a trend toward less PM-1 positive neutrophil compared with the 316L equivalent stent. Conclusion Despite having identical scaffold characteristics regarding geometrical design, Magmaris had significantly less thrombogenicity and inflammatory cell deposition compared with the equivalent 316L stainless steel stent in a porcine arteriovenous shunt model. These data suggest resorbable magnesium scaffolds may have inherent properties that reduce adhesion of platelets and inflammatory cells.

  12. Quantitation of thrombogenicity of hemodialyzer with technetium-99m and indium-111 labeled platelets

    Energy Technology Data Exchange (ETDEWEB)

    Dewanjee, M.K.; Kapadvanjwala, Mansoor; Ruzius, Kees; Serafini, A.N.; Zilleruelo, G.E.; Sfakianakis, G.N. (Miami Univ., FL (United States). School of Medicine Althin CD-Medical Inc., Miami Lakes, FL (United States))

    1993-07-01

    The platelet thromobogenicity of a hemodialyzer was quantified with [sup 99m]Tc- and [sup 111]In-labeled platelets. The platelets collected from blood of Beagle dogs, Yorkshire pigs and human volunteers were labeled with [sup 111]in-tropolone (detergent-free) and [sup 99m]Tc-HMPAO. Hemodialysis was performed with a hollow-fiber dialyzer (HFD) in a flow-loop, the temperature of which was maintained at 37[sup o]C, with flow-rates of 7, 150 and 270 mL/min; after dialysis, the HFD radioactivity was measured with an ionization chamber and imaged with a [gamma]-camera. The radioactivity of samples of hollow-fibers taken from the top, middle and bottom of the dialyzer was determined with a [gamma]-counter. The mean values of hemodialyzer-adherent platelet radioactivity were calculated for both radionuclides. The canine platelets were found to be more thrombogenic than porcine and human platelets. The adhesivity of porcine platelets to the biomaterial (cellulose-acetate) of the dialyzer approximated that of human platelets. The [sup 99m]Tc label underestimated the thrombus formation (P < 0.01 ). The dynamic processes of thrombosis and embolization from the hemodialyzer resulted in the large standard deviations around the mean values of the adherent thrombus. In spite of this limitation of the dynamic pathology, the quantitation of comparative throbogenicity with [sup 111]In- and [sup 99m]Tc-labeled platelets suggests that both radionuclides could be used for measurement of device-induced thrombogenicity and may provide an estimation of prosthesis-induced thrombogenicity of human platelets from animal studies. (Author).

  13. IEEE Std 382-1985: IEEE standard for qualification of actuators for power operated valve assemblies with safety-related functions for nuclear power plants

    International Nuclear Information System (INIS)

    Anon.

    1992-01-01

    This standard describes the qualification of all types of power-driven valve actuators, including damper actuators, for safety-related functions in nuclear power generating stations. This standard may also be used to separately qualify actuator components. This standard establishes the minimum requirements for, and guidance regarding, the methods and procedures for qualification of power-driven valve actuators with safety-related functions Part I describes the qualification process. Part II describes the standard qualification cases and their environmental parameters for the usual locations of safety-related equipment in a nuclear generating station. Part III describes the qualification tests outlined in 6.3.3

  14. Transcatheter Mitral Valve-in-Ring Implantation

    LENUS (Irish Health Repository)

    Tanner, RE

    2018-05-01

    Failed surgical mitral valve repair using an annuloplasty ring has traditionally been treated with surgical valve replacement or repair1. For patients at high risk for repeat open heart surgery, placement of a trans-catheter aortic valve (i.e., TAVI valve) within the mitral ring (i.e., Mitral-Valve-in-Ring, MViR) has emerged as a novel alternative treatment strategy2-5 . We describe our experience of a failed mitral valve repair that was successfully treated with a TAVI valve delivered via the trans-septal approach, and summarise the data relating to this emerging treatment strategy.

  15. The Development of the Safety Related Valve Class 1E Electrical Motor, the Target and the Results

    International Nuclear Information System (INIS)

    Saban, I.; Grgic, D.; Fancev, T.; Flegar, Lj.; Novosel, N.

    1996-01-01

    The development of the safety related valves class 1E electric motor is described. The design implemented in order to satisfy the 1E requirements, and a way in which related 1E standards are addressed, are shown. The development was realized in three stages. In the first stage eight motorettes were made and the insulation system was tested. In the second stage the motor was produced in accordance with producer's prototype QA program. In the third stage part of the testing of the produced motor was made. The results of the testing, finished until now, show that produced motor, as well as similarly produced motors, is able to perform its safety function in the design bases accident conditions as requested by class 1E requirements. The rest of the testing (LOCA test) can be made on the same or similar motor in the future. (author)

  16. Mitral Valve Disease

    Science.gov (United States)

    ... for mitral valve replacement—mechanical valves (metal) or biological valves (tissue). The principal advantage of mechanical valves ... small risk of stroke due to blood clotting. Biological valves usually are made from animal tissue. Biological ...

  17. Platelet thrombosis in cardiac-valve prostheses

    International Nuclear Information System (INIS)

    Dewanjee, M.K.

    1989-01-01

    The contribution of platelets and clotting factors in thrombosis on cardiovascular prostheses had been quantified with several tracers. Thrombus formation in vivo could be measured semiquantitatively in animal models and patients with indium-111, Technetium-99m labeled platelets, iodine-123, iodine-131 labeled fibrinogen, and In-111 and Tc-99m labeled antibody to the fibrinogen-receptor on the platelet- membrane, or fibrin. The early studies demonstrated that certain platelet-inhibitors, e.g. sulfinpyrazone, aspirin or aspirin- persantine increased platelet survival time with mechanical valves implanted in the baboon model and patients. Thrombus localization by imaging is possible for large thrombus on thrombogenic surface of prosthesis in the acute phase. The majority of thrombus was found in the sewing ring (Dacron) in the acute phase in both the mechanical and tissue valves. The amount of retained thrombus in both mechanical and tissue valves in our one-day study in the dog model was similar (< 1% if injected In-111 platelets = 5 billion platelets). As the fibrous ingrowth covered the sewing ring, the thrombus formation decreased significantly. Only a small amount of thrombus was found on the leaflets at one month in both the dog and calf models. 38 refs., 9 figs., 5 tabs

  18. Platelet thrombosis in cardiac-valve prostheses

    Energy Technology Data Exchange (ETDEWEB)

    Dewanjee, M.K.

    1989-01-01

    The contribution of platelets and clotting factors in thrombosis on cardiovascular prostheses had been quantified with several tracers. Thrombus formation in vivo could be measured semiquantitatively in animal models and patients with indium-111, Technetium-99m labeled platelets, iodine-123, iodine-131 labeled fibrinogen, and In-111 and Tc-99m labeled antibody to the fibrinogen-receptor on the platelet- membrane, or fibrin. The early studies demonstrated that certain platelet-inhibitors, e.g. sulfinpyrazone, aspirin or aspirin- persantine increased platelet survival time with mechanical valves implanted in the baboon model and patients. Thrombus localization by imaging is possible for large thrombus on thrombogenic surface of prosthesis in the acute phase. The majority of thrombus was found in the sewing ring (Dacron) in the acute phase in both the mechanical and tissue valves. The amount of retained thrombus in both mechanical and tissue valves in our one-day study in the dog model was similar (< 1% if injected In-111 platelets = 5 billion platelets). As the fibrous ingrowth covered the sewing ring, the thrombus formation decreased significantly. Only a small amount of thrombus was found on the leaflets at one month in both the dog and calf models. 38 refs., 9 figs., 5 tabs.

  19. The involvement of multiple thrombogenic and atherogenic markers in premature coronary artery disease

    Directory of Open Access Journals (Sweden)

    Antonio P. Mansur

    2013-12-01

    Full Text Available OBJECTIVE: To examine the association of atherogenic and thrombogenic markers and lymphotoxin-alfa gene mutations with the risk of premature coronary disease. METHODS: This cross-sectional, case-control, age-adjusted study was conducted in 336 patients with premature coronary disease (50% luminal reduction or a previous myocardial infarction. The laboratory data evaluated included thrombogenic factors (fibrinogen, protein C, protein S, and antithrombin III, atherogenic factors (glucose and lipid profiles, lipoprotein(a, and apolipoproteins AI and B, and lymphotoxin-alfa mutations. Genetic variability of lymphotoxin-alfa was determined by polymerase chain reaction analysis. RESULTS: Coronary disease patients exhibited lower concentrations of HDL-cholesterol and higher levels of glucose, lipoprotein(a, and protein S. The frequencies of AA, AG, and GG lymphotoxin-alfa mutation genotypes were 55.0%, 37.6%, and 7.4% for controls and 42.7%, 46.0%, and 11.3% for coronary disease patients (p = 0.02, respectively. Smoking, dyslipidemia, family history, and lipoprotein(a and lymphotoxin-alfa mutations in men were independent variables associated with coronary disease. The area under the curve (C-statistic increased from 0.779 to 0.802 (p<0.05 with the inclusion of lipoprotein(a and lymphotoxin-alfa mutations in the set of conventional risk factors. CONCLUSIONS: The inclusion of lipoprotein(a and lymphotoxin-alfa mutations in the set of conventional risk factors showed an additive but small increase in the risk prediction of premature coronary disease.

  20. Relative Expression of PBMC MicroRNA-133a Analysis in Patients Receiving Warfarin After Mechanical Heart Valve Replacement

    Science.gov (United States)

    Kabiri Rad, Hamid; Mazaheri, Mahta; Dehghani Firozabadi, Ali

    Background: MicroRNAs (miRNAs) are implicated in various biological processes including anticoagulation. However, the modulation of miRNA by pharmacological intervention such as warfarin treatment in patients receiving warfarin has not been disclosed yet. The aim of this study work was to assess the effect of warfarin drug on expression level of mir-133a-3p in patients with mechanical heart valve replacement. Methods: In this research, the expression level of miRNA-133a-3p was analyzed in Peripheral Blood Mononuclear Cells (PBMCs) from mechanical valve replacement patients who had received warfarin for at least 3 months continuously. Quantitative RT-PCR method was used for this assay. Results: Our findings indicated a significant diffrence between the rate of miR-133a-3p expression in individuals receiving warfarin and the control group (p<0.01). There was also a statistically significant difference in miR-133a-3p expression in patients with different ages (p<0.05) suggesting that the rate of miR-133a-3p expression in persons receiving warfarin is related to age. However, other variables like warfarin dose, International Normalized Ratio (INR), gender, and Body Mass Index (BMI) were not significantly effective on the miR-133a-3p experssion rate in individuals receving warfarin. Conclusion: Based on our results, it can be concluded that miR-133a-3p is involved in the coagulation pathway. The recent result indicates that warfarin affects the expression of miR-133a. This expression may be potentially important for treatment by anticoagulants. Awareness of the time course of miRNA expression profile can improve efficiency of response to warfarin. PMID:29296264

  1. Age-related changes in aortic 3D blood flow velocities and wall shear stress: Implications for the identification of altered hemodynamics in patients with aortic valve disease

    NARCIS (Netherlands)

    van Ooij, Pim; Garcia, Julio; Potters, Wouter V.; Malaisrie, S. Chris; Collins, Jeremy D.; Carr, James C.; Markl, Michael; Barker, Alex J.

    2016-01-01

    To investigate age-related changes in peak systolic aortic 3D velocity and wall shear stress (WSS) in healthy controls and to investigate the importance of age-matching for 3D mapping of abnormal aortic hemodynamics in bicuspid aortic valve disease (BAV). 4D flow MRI (fields strengths = 1.5-3T;

  2. Bistable fluidic valve is electrically switched

    Science.gov (United States)

    Fiet, O.; Salvinski, R. J.

    1970-01-01

    Bistable control valve is selectively switched by direct application of an electrical field to divert fluid from one output channel to another. Valve is inexpensive, has no moving parts, and operates on fluids which are relatively poor electrical conductors.

  3. Fluid mechanics of heart valves.

    Science.gov (United States)

    Yoganathan, Ajit P; He, Zhaoming; Casey Jones, S

    2004-01-01

    Valvular heart disease is a life-threatening disease that afflicts millions of people worldwide and leads to approximately 250,000 valve repairs and/or replacements each year. Malfunction of a native valve impairs its efficient fluid mechanic/hemodynamic performance. Artificial heart valves have been used since 1960 to replace diseased native valves and have saved millions of lives. Unfortunately, despite four decades of use, these devices are less than ideal and lead to many complications. Many of these complications/problems are directly related to the fluid mechanics associated with the various mechanical and bioprosthetic valve designs. This review focuses on the state-of-the-art experimental and computational fluid mechanics of native and prosthetic heart valves in current clinical use. The fluid dynamic performance characteristics of caged-ball, tilting-disc, bileaflet mechanical valves and porcine and pericardial stented and nonstented bioprostheic valves are reviewed. Other issues related to heart valve performance, such as biomaterials, solid mechanics, tissue mechanics, and durability, are not addressed in this review.

  4. Double-disc gate valve

    International Nuclear Information System (INIS)

    Wheatley, S.J.

    1979-01-01

    The invention relates to an improvement in a conventional double-disc gate valve having a vertically movable gate assembly including a wedge, spreaders slidably engaged therewith, a valve disc carried by the spreaders. When the gate assembly is lowered to a selected point in the valve casing, the valve discs are moved transversely outward to close inlet and outlet ports in the casing. The valve includes hold-down means for guiding the disc-and-spreader assemblies as they are moved transversely outward and inward. If such valves are operated at relatively high differential pressures, they sometimes jam during opening. Such jamming has been a problem for many years in gate valves used in gaseous diffusion plants for the separation of uranium isotopes. The invention is based on the finding that the above-mentioned jamming results when the outlet disc tilts about its horizontal axis in a certain way during opening of the valve. In accordance with the invention, tilting of the outlet disc is maintained at a tolerable value by providing the disc with a rigid downwardly extending member and by providing the casing with a stop for limiting inward arcuate movement of the member to a preselected value during opening of the valve

  5. Comparison of Acute Thrombogenicity for Metallic and Polymeric Bioabsorbable Scaffolds: Magmaris Versus Absorb in a Porcine Arteriovenous Shunt Model.

    Science.gov (United States)

    Waksman, Ron; Lipinski, Michael J; Acampado, Eduardo; Cheng, Qi; Adams, Lila; Torii, Sho; Gai, Jiaxiang; Torguson, Rebecca; Hellinga, David M; Westman, Peter C; Joner, Michael; Zumstein, Philine; Kolodgie, Frank D; Virmani, Renu

    2017-08-01

    A comparison in acute thrombogenicity between the Magmaris sirolimus-eluting bioabsorbable magnesium scaffold and the Absorb bioresorbable vascular scaffold has not been performed. This study assessed acute thrombogenicity of Magmaris compared with Absorb and the Orsiro hybrid drug-eluting stent in a porcine arteriovenous shunt model. An ex vivo porcine carotid jugular arteriovenous shunt was established and connected to SYLGARD tubing containing the Magmaris, Absorb, and Orsiro scaffolds/stents and allowed to run in the shunt for a maximum of 1 hour. Twelve shunts (2 shunt runs per pig) were run comparing the 3 scaffolds in alternating order. Nested generalized linear mixed models were used to compare variables between scaffold groups while adjusting for variability between shunt runs. Confocal fluorescent microscopy costaining CD61/CD42b demonstrated that both Magmaris (3.0%) and Orsiro (4.6%) had less platelet coverage of the total scaffold compared with Absorb (21.8%). Scanning electron microscopy demonstrated significantly less thrombus deposition to Magmaris as a percentage of the total scaffold compared with Absorb (5.0% versus 16.1%, P =0.02). Magmaris had significantly less PM-1-positive neutrophil and CD14-positive monocyte adherence compared with both Orsiro and Absorb. Orsiro had significantly less monocyte deposition compared with Absorb. Despite a similar scaffold strut thickness, the Magmaris sirolimus-eluting bioabsorbable magnesium scaffold was significantly less thrombogenic compared with the Absorb bioresorbable vascular scaffold in an ex vivo porcine arteriovenous shunt model. Further studies are needed to determine whether the reduced thrombogenicity of Magmaris will result in reductions in major cardiovascular events. © 2017 American Heart Association, Inc.

  6. 3D velocity field characterization of prosthetic heart valve with two different valve testers by means of stereo-PIV.

    Science.gov (United States)

    D'Avenio, Giuseppe; Grigioni, Mauro; Daniele, Carla; Morbiducci, Umberto; Hamilton, Kathrin

    2015-01-01

    Prosthetic heart valves can be associated to mechanical loading of blood, potentially linked to complications (hemolysis and thrombogenicity) which can be clinically relevant. In order to test such devices in pulsatile mode, pulse duplicators (PDs) have been designed and built according to different concepts. This study was carried out to compare anemometric measurements made on the same prosthetic device, with two widely used PDs. The valve (a 27-mm bileaflet valve) was mounted in the aortic section of the PD. The Sheffield University PD and the RWTH Aachen PD were selected as physical models of the circulation. These two PDs differ mainly in the vertical vs horizontal realization, and in the ventricular section, which in the RWTH PD allows for storage of potential energy in the elastic walls of the ventricle. A glassblown aorta, realized according to the geometric data of the same anatomical district in healthy individuals, was positioned downstream of the valve, obtaining 1:1 geometric similarity conditions. A NaI-glycerol-water solution of suitable kinematic viscosity and, at the same time, the proper refractive index, was selected. The flow field downstream of the valve was measured by means of the stereo-PIV (Particle Image Velocimetry) technique, capable of providing the complete 3D velocity field as well as the entire Reynolds stress tensor. The measurements were carried out at the plane intersecting the valve axis. A three-jet profile was clearly found in the plane crossing the leaflets, with both PDs. The extent of the typical recirculation zone in the Valsalva sinus was much larger in the RWTH PD, on account of the different duration of the swirling motion in the ventricular chamber, caused by the elasticity of the ventricle and its geometry. The comparison of the hemodynamical behaviour of the same bileaflet valve tested in two PDs demonstrated the role of the mock loop in affecting the valve performance.

  7. Geometric Optimization for Non-Thrombogenicity of a Centrifugal Blood Pump through Flow Visualization

    Science.gov (United States)

    Toyoda, Masahiro; Nishida, Masahiro; Maruyama, Osamu; Yamane, Takashi; Tsutsui, Tatsuo; Sankai, Yoshiyuki

    A monopivot centrifugal blood pump, whose impeller is supported with a pivot bearing and a passive magnetic bearing, is under development for implantable artificial heart. The hemolysis level is less than that of commercial centrifugal pumps and the pump size is as small as 160 mL in volume. To solve a problem of thrombus caused by fluid dynamics, flow visualization experiments and animal experiments have been undertaken. For flow visualization a three-fold scale-up model, high-speed video system, and particle tracking velocimetry software were used. To verify non-thrombogenicity one-week animal experiments were conducted with sheep. The initially observed thrombus around the pivot was removed through unifying the separate washout holes to a small centered hole to induce high shear around the pivot. It was found that the thrombus contours corresponded to the shear rate of 300s-1 for red thrombus and 1300-1700s-1 for white thrombus, respectively. Thus flow visualization technique was found to be a useful tool to predict thrombus location.

  8. Geometric Optimization for Non-Thrombogenicity of a Centrifugal Blood Pump through Computational Fluid Dynamic Analysis

    Science.gov (United States)

    Nishida, Masahiro; Yamane, Takashi

    A monopivot magnetic suspension blood pump has been developed in our laboratory. The flow patterns within the pump should be carefully examined in order to prevent thrombogenesis, especially around the pivot bearing. Therefore, the effects of the pump geometry on the local flow were analyzed using computational fluid dynamics together with the experimental flow visualization. The engineering goal was to reduce the area of stagnation around the pivot in order to prevent thrombus formation. As a result, the stagnation area and the flow rate through the washout holes were found to be highly affected by the size and geometry of the washout holes. Secondary flow was revealed to form a jet-like wash against the pivot, thus preventing thrombus formation. The flow rate through the washout holes was estimated to be up to one fifth of the pump flow rate, depending on the cross-sectional areas of the washout holes. Furthermore, an anti-thrombogenic effect was attained by removing a small gap between the male and female pivots.

  9. Targeting Heparin to Collagen within Extracellular Matrix Significantly Reduces Thrombogenicity and Improves Endothelialization of Decellularized Tissues.

    Science.gov (United States)

    Jiang, Bin; Suen, Rachel; Wertheim, Jason A; Ameer, Guillermo A

    2016-12-12

    Thrombosis within small-diameter vascular grafts limits the development of bioartificial, engineered vascular conduits, especially those derived from extracellular matrix (ECM). Here we describe an easy-to-implement strategy to chemically modify vascular ECM by covalently linking a collagen binding peptide (CBP) to heparin to form a heparin derivative (CBP-heparin) that selectively binds a subset of collagens. Modification of ECM with CBP-heparin leads to increased deposition of functional heparin (by ∼7.2-fold measured by glycosaminoglycan composition) and a corresponding reduction in platelet binding (>70%) and whole blood clotting (>80%) onto the ECM. Furthermore, addition of CBP-heparin to the ECM stabilizes long-term endothelial cell attachment to the lumen of ECM-derived vascular conduits, potentially through recruitment of heparin-binding growth factors that ultimately improve the durability of endothelialization in vitro. Overall, our findings provide a simple yet effective method to increase deposition of functional heparin on the surface of ECM-based vascular grafts and thereby minimize thrombogenicity of decellularized tissue, overcoming a significant challenge in tissue engineering of bioartificial vessels and vascularized organs.

  10. Poppet valve tester

    Science.gov (United States)

    Tellier, G. F.

    1973-01-01

    Tester investigates fundamental factors affecting cyclic life and sealing performance of valve seats and poppets. Tester provides for varying impact loading of poppet against seat and rate of cycling, and controls amount and type of relative motion between sealing faces of seat and poppet. Relative motion between seat and poppet can be varied in three modes.

  11. Antiplatelet and anticoagulation regimen in patients with mechanical valve undergoing PCI - State-of-the-art review.

    Science.gov (United States)

    Gajanana, Deepakraj; Rogers, Toby; Iantorno, Micaela; Buchanan, Kyle D; Ben-Dor, Itsik; Pichard, Augusto D; Satler, Lowell F; Torguson, Rebecca; Okubagzi, Petros G; Waksman, Ron

    2018-04-02

    A common clinical dilemma regarding treatment of patients with a mechanical valve is the need for concomitant antiplatelet therapy for a variety of reasons, referred to as triple therapy. Triple therapy is when a patient is prescribed aspirin, a P2Y12 antagonist, and an oral anticoagulant. Based on the totality of the available evidence, best practice in 2017 for patients with mechanical valves undergoing percutaneous coronary intervention (PCI) is unclear. Furthermore, the optimal duration of dual antiplatelet therapy after PCI is evolving. With better valve designs that are less thrombogenic, the thromboembolic risks can be reduced at a lower international normalized ratio target, thus decreasing the bleeding risk. This review will offer an in-depth survey of current guidelines, current evidence, suggested approach for PCI in this cohort, and future studies regarding mechanical valve patients undergoing PCI. Copyright © 2017 Elsevier B.V. All rights reserved.

  12. Aortic valve insufficiency in the teenager and young adult: the role of prosthetic valve replacement.

    Science.gov (United States)

    Bradley, Scott M

    2013-10-01

    The contents of this article were presented in the session "Aortic insufficiency in the teenager" at the congenital parallel symposium of the 2013 Society of Thoracic Surgeons (STS) annual meeting. The accompanying articles detail the approaches of aortic valve repair and the Ross procedure.(1,2) The current article focuses on prosthetic valve replacement. For many young patients requiring aortic valve surgery, either aortic valve repair or a Ross procedure provides a good option. The advantages include avoidance of anticoagulation and potential for growth. In other patients, a prosthetic valve is an appropriate alternative. This article discusses the current state of knowledge regarding mechanical and bioprosthetic valve prostheses and their specific advantages relative to valve repair or a Ross procedure. In current practice, young patients requiring aortic valve surgery frequently undergo valve replacement with a prosthetic valve. In STS adult cardiac database, among patients ≤30 years of age undergoing aortic valve surgery, 34% had placement of a mechanical valve, 51% had placement of a bioprosthetic valve, 9% had aortic valve repair, and 2% had a Ross procedure. In the STS congenital database, among patients 12 to 30 years of age undergoing aortic valve surgery, 21% had placement of a mechanical valve, 18% had placement of a bioprosthetic valve, 30% had aortic valve repair, and 24% had a Ross procedure. In the future, the balance among these options may be altered by design improvements in prosthetic valves, alternatives to warfarin, the development of new patch materials for valve repair, and techniques to avoid Ross autograft failure.

  13. Diagnostic accuracy of electrocardiographic P wave related parameters in the assessment of left atrial size in dogs with degenerative mitral valve disease.

    Science.gov (United States)

    Soto-Bustos, Ángel; Caro-Vadillo, Alicia; Martínez-DE-Merlo, Elena; Alonso-Alegre, Elisa González

    2017-10-07

    The purpose of this research was to compare the accuracy of newly described P wave-related parameters (P wave area, Macruz index and mean electrical axis) with classical P wave-related parameters (voltage and duration of P wave) for the assessment of left atrial (LA) size in dogs with degenerative mitral valve disease. One hundred forty-six dogs (37 healthy control dogs and 109 dogs with degenerative mitral valve disease) were prospectively studied. Two-dimensional echocardiography examinations and a 6-lead ECG were performed prospectively in all dogs. Echocardiography parameters, including determination of the ratios LA diameter/aortic root diameter and LA area/aortic root area, were compared to P wave-related parameters: P wave area, Macruz index, mean electrical axis voltage and duration of P wave. The results showed that P wave-related parameters (classical and newly described) had low sensitivity (range=52.3 to 77%; median=60%) and low to moderate specificity (range=47.2 to 82.5%; median 56.3%) for the prediction of left atrial enlargement. The areas under the curve of P wave-related parameters were moderate to low due to poor sensitivity. In conclusion, newly P wave-related parameters do not increase the diagnostic capacity of ECG as a predictor of left atrial enlargement in dogs with degenerative mitral valve disease.

  14. Beyond adding years to life: health-related quality-of-life and functional outcomes in patients with severe aortic valve stenosis at high surgical risk undergoing transcatheter aortic valve replacement.

    Science.gov (United States)

    Deutsch, Marcus-Andre; Bleiziffer, Sabine; Elhmidi, Yacine; Piazza, Nicolo; Voss, Bernhard; Lange, Ruediger; Krane, Markus

    2013-11-01

    Aortic valve stenosis (AVS) is the most frequent acquired valvular heart disease in western industrialized countries and its prevalence considerably increases with age. Once becoming symptomatic severe AVS has a very poor prognosis. Progressive and rapid symptom deterioration leads to an impairment of functional status and compromised healthrelated quality-of-life (HrQoL) simultaneously. Until recently, surgical aortic valve replacement (SAVR) has been the only effective treatment option for improving symptoms and prolonging survival. Transcatheter aortic valve replacement (TAVR) emerged as an alternative treatment modality for those patients with severe symptomatic AVS in whom the risk for SAVR is considered prohibitive or too high. TAVR has gained clinical acceptance with almost startling rapidity and has even quickly become the standard of care for the treatment of appropriately selected individuals with inoperable AVS during recent years. Typically, patients currently referred for and treated by TAVR are elderly with a concomitant variable spectrum of multiple comorbidities, disabilities and limited life expectancy. Beyond mortality and morbidity, the assessment of HrQoL is of paramount importance not only to guide patient-centered clinical decision-making but also to judge this new treatment modality. As per current evidence, TAVR significantly improves HrQoL in high-surgical risk patients with severe AVS with sustained effects up to two years when compared with optimal medical care and demonstrates comparable benefits relative to SAVR. Along with a provision of a detailed overview of the current literature regarding functional and HrQoL outcomes in patients undergoing TAVR, this review article addresses specific considerations of the HrQoL aspect in the elderly patient and finally outlines the implications of HrQoL outcomes for medico-economic deliberations.

  15. Piezoelectric valve

    Science.gov (United States)

    Petrenko, Serhiy Fedorovich

    2013-01-15

    A motorized valve has a housing having an inlet and an outlet to be connected to a pipeline, a saddle connected with the housing, a turn plug having a rod, the turn plug cooperating with the saddle, and a drive for turning the valve body and formed as a piezoelectric drive, the piezoelectric drive including a piezoelectric generator of radially directed standing acoustic waves, which is connected with the housing and is connectable with a pulse current source, and a rotor operatively connected with the piezoelectric generator and kinematically connected with the rod of the turn plug so as to turn the turn plug when the rotor is actuated by the piezoelectric generator.

  16. Aging and service wear of spring-loaded pressure relief valves used in safety-related systems at nuclear power plants

    International Nuclear Information System (INIS)

    Staunton, R.H.; Cox, D.F.

    1995-03-01

    Spring-loaded pressure relief valves (PRVS) are used in some safety-related applications at nuclear power plants. In general, they are used in systems where, during accidents, pressures may rise to levels where pressure safety relief is required for protection of personnel, system piping, and components. This report documents a study of PRV aging and considers the severity and causes of service wear and how it is discovered and corrected in various systems, valve sizes, etc. Provided in this report are results of the examination of the recorded failures and identification of trends and relationships/correlations in the failures when all failure-related parameters are considered. Components that comprise a typical PRV, how those components fail, when they fail, and the current testing frequencies and methods are also presented in detail

  17. Aging and service wear of spring-loaded pressure relief valves used in safety-related systems at nuclear power plants

    Energy Technology Data Exchange (ETDEWEB)

    Staunton, R.H.; Cox, D.F. [Oak Ridge National Lab., TN (United States)

    1995-03-01

    Spring-loaded pressure relief valves (PRVS) are used in some safety-related applications at nuclear power plants. In general, they are used in systems where, during accidents, pressures may rise to levels where pressure safety relief is required for protection of personnel, system piping, and components. This report documents a study of PRV aging and considers the severity and causes of service wear and how it is discovered and corrected in various systems, valve sizes, etc. Provided in this report are results of the examination of the recorded failures and identification of trends and relationships/correlations in the failures when all failure-related parameters are considered. Components that comprise a typical PRV, how those components fail, when they fail, and the current testing frequencies and methods are also presented in detail.

  18. Echocardiographic evaluation of heart valve prosthetic dysfunction

    Directory of Open Access Journals (Sweden)

    Yuriy Ivaniv

    2018-02-01

    Full Text Available Patients with replaced heart valve submitted to echocardiographic examination may have symptoms related either to valvular malfunction or ventricular dysfunction from different causes. Clinical examination is not reliable in a prosthetic valve evaluation and the main information regarding its function could be obtained using different cardiac ultrasound modalities. This review provides a description of echocardiographic and Doppler techniques useful in evaluation of prosthetic heart valves. For the interpretation of echocardiography there is a need in special knowledge of prosthesis types and possible reasons of prosthetic function deterioration. Echocardiography allows to reveal valve thrombosis, pannus formation, vegetation and such complications of infective endocarditis as valve ring abscess or dehiscence. Transthoracic echocardiography requires different section plane angles and unconventional views. Transesophageal echocardiography is more often used than in native valve examination due to better visualization of prosthetic valve structure and function. Three-dimensional echocardiography could provide more detailed visual information especially in the assessment of paravalvular regurgitation or valve obstruction.

  19. A symmetric safety valve

    International Nuclear Information System (INIS)

    Burtraw, Dallas; Palmer, Karen; Kahn, Danny

    2010-01-01

    How to set policy in the presence of uncertainty has been central in debates over climate policy. Concern about costs has motivated the proposal for a cap-and-trade program for carbon dioxide, with a 'safety valve' that would mitigate against spikes in the cost of emission reductions by introducing additional emission allowances into the market when marginal costs rise above the specified allowance price level. We find two significant problems, both stemming from the asymmetry of an instrument that mitigates only against a price increase. One is that most important examples of price volatility in cap-and-trade programs have occurred not when prices spiked, but instead when allowance prices collapsed. Second, a single-sided safety valve may have unintended consequences for investment. We illustrate that a symmetric safety valve provides environmental and welfare improvements relative to the conventional one-sided approach.

  20. Guidelines for valves in tritium service

    International Nuclear Information System (INIS)

    Weaver, W.W.

    1994-01-01

    Some undesirable practices and misapplications that caused valve-related failures are examined, and future courses of action are recommended to avoid repetition of these events. Desirable valve characteristics and practices that should be considered when selecting valves for use in tritium service are also discussed. Supporting logic for the desirability of these features is presented by discussing the mechanisms of valve degradation followed by examples of related events. Desirable valve and system features and operational actions are grouped into two categories: strongly recommended and recommended. 13 refs., 1 fig

  1. Heart valve surgery - discharge

    Science.gov (United States)

    ... ACC guideline for the management of patients with valvular heart disease: executive summary: a report of the American College ... Editorial team. Related MedlinePlus Health Topics Heart Surgery Heart Valve Diseases Browse the Encyclopedia A.D.A.M., Inc. ...

  2. Improved thrombogenicity on oxygen etched Ti6Al4V surfaces

    Energy Technology Data Exchange (ETDEWEB)

    Riedel, Nicholas A. [Department of Mechanical Engineering, Colorado State University, Fort Collins, CO 80523 (United States); Smith, Barbara S. [School of Biomedical Engineering, Colorado State University, Fort Collins, CO 80523 (United States); Williams, John D. [Department of Mechanical Engineering, Colorado State University, Fort Collins, CO 80523 (United States); Popat, Ketul C., E-mail: ketul.popat@colostate.edu [Department of Mechanical Engineering, Colorado State University, Fort Collins, CO 80523 (United States); School of Biomedical Engineering, Colorado State University, Fort Collins, CO 80523 (United States)

    2012-07-01

    Thrombus formation on blood contacting biomaterials continues to be a key factor in initiating a critical mode of failure in implantable devices, requiring immediate attention. In the interest of evaluating a solution for one of the most widely used biomaterials, titanium and its alloys, this study focuses on the use of a novel surface oxidation treatment to improve the blood compatibility. This study examines the possibility of using oblique angle ion etching to produce a high quality oxide layer that enhances blood compatibility on medical grade titanium alloy Ti6Al4V. An X-ray photoelectron spectroscopy (XPS) analysis of these oxygen-rich surfaces confirmed the presence of TiO{sub 2} peaks and also indicated increased surface oxidation as well as a reduction in surface defects. After 2 h of contact with whole human plasma, the oxygen etched substrates demonstrated a reduction in both platelet adhesion and activation as compared to bare titanium substrates. The whole blood clotting behavior was evaluated for up to 45 min, showing a significant decrease in clot formation on oxygen etched substrates. Finally, a bicinchoninic acid (BCA) total protein assay and XPS were used to evaluate the degree of key blood serum protein (fibrinogen, albumin, immunoglobulin G) adsorption on the substrates. The results showed similar protein levels for both the oxygen etched and control substrates. These results indicate that oblique angle oxygen etching may be a promising method to increase the thrombogenicity of Ti6Al4V. - Highlights: Black-Right-Pointing-Pointer Oblique angle oxygen ion etching creates a high quality, uniform oxide surface. Black-Right-Pointing-Pointer Oxygen etched substrates showed fewer adhered platelets. Black-Right-Pointing-Pointer Platelet activation was reduced by the improved oxide surface. Black-Right-Pointing-Pointer Oxygen etched substrates exhibited increased whole blood clotting times. Black-Right-Pointing-Pointer Although clotting reductions were

  3. Improved thrombogenicity on oxygen etched Ti6Al4V surfaces

    International Nuclear Information System (INIS)

    Riedel, Nicholas A.; Smith, Barbara S.; Williams, John D.; Popat, Ketul C.

    2012-01-01

    Thrombus formation on blood contacting biomaterials continues to be a key factor in initiating a critical mode of failure in implantable devices, requiring immediate attention. In the interest of evaluating a solution for one of the most widely used biomaterials, titanium and its alloys, this study focuses on the use of a novel surface oxidation treatment to improve the blood compatibility. This study examines the possibility of using oblique angle ion etching to produce a high quality oxide layer that enhances blood compatibility on medical grade titanium alloy Ti6Al4V. An X-ray photoelectron spectroscopy (XPS) analysis of these oxygen-rich surfaces confirmed the presence of TiO 2 peaks and also indicated increased surface oxidation as well as a reduction in surface defects. After 2 h of contact with whole human plasma, the oxygen etched substrates demonstrated a reduction in both platelet adhesion and activation as compared to bare titanium substrates. The whole blood clotting behavior was evaluated for up to 45 min, showing a significant decrease in clot formation on oxygen etched substrates. Finally, a bicinchoninic acid (BCA) total protein assay and XPS were used to evaluate the degree of key blood serum protein (fibrinogen, albumin, immunoglobulin G) adsorption on the substrates. The results showed similar protein levels for both the oxygen etched and control substrates. These results indicate that oblique angle oxygen etching may be a promising method to increase the thrombogenicity of Ti6Al4V. - Highlights: ►Oblique angle oxygen ion etching creates a high quality, uniform oxide surface. ►Oxygen etched substrates showed fewer adhered platelets. ►Platelet activation was reduced by the improved oxide surface. ►Oxygen etched substrates exhibited increased whole blood clotting times. ►Although clotting reductions were seen, protein adsorption remained similar.

  4. Thermal overload protection for electric motors on safety-related motor-operated valves: Generic Issue II.E.6.1

    International Nuclear Information System (INIS)

    Rothberg, O.

    1988-06-01

    NRC regulatory positions, as stated in Regulatory Guide 1.106, Revision 1, have been identified by the Office for Analysis and Evaluation of Operational Data (AEOD) as potential contributors to valve motor burnout. AEOD is particularly concerned about the allowed policy of bypassing thermal overload devices during normal or accident conditions. Regulatory Guide 1.106 favors compromising the function of thermal overload devices in favor of completing the safety-related action of valves. The purpose of this study was to determine if the guidance contained in Regulatory Guide 1.106 is appropriate and, if not, to recommend the necessary changes. This report describes thermal overload devices commonly used to protect safety-related valve operator motors. The regulatory guidelines stated in Regulatory Guide 1.106 along with the limitations of thermal overload protection are discussed. Supplements and alternatives to thermal overload protection are also described. Findings and conclusions of several AEOD reports are discussed. Information obtained from the standard review plan, standard technical specifications, technical specifications from representative plants, and several papers are cited

  5. Computational evaluation of the thrombogenic potential of a hollow-fiber oxygenator with integrated heat exchanger during extracorporeal circulation.

    Science.gov (United States)

    Pelosi, Alessandra; Sheriff, Jawaad; Stevanella, Marco; Fiore, Gianfranco B; Bluestein, Danny; Redaelli, Alberto

    2014-04-01

    The onset of thromboembolic phenomena in blood oxygenators, even in the presence of adequate anticoagulant strategies, is a relevant concern during extracorporeal circulation (ECC). For this reason, the evaluation of the thrombogenic potential associated with extracorporeal membrane oxygenators should play a critical role into the preclinical design process of these devices. This study extends the use of computational fluid dynamics simulations to guide the hemodynamic design optimization of oxygenators and evaluate their thrombogenic potential during ECC. The computational analysis accounted for both macro- (i.e., vortex formation) and micro-scale (i.e., flow-induced platelet activation) phenomena affecting the performances of a hollow-fiber membrane oxygenator with integrated heat exchanger. A multiscale Lagrangian approach was adopted to infer the trajectory and loading history experienced by platelet-like particles in the entire device and in a repetitive subunit of the fiber bundles. The loading history was incorporated into a damage accumulation model in order to estimate the platelet activation state (PAS) associated with repeated passes of the blood within the device. Our results highlighted the presence of blood stagnation areas in the inlet section that significantly increased the platelet activation levels in particles remaining trapped in this region. The order of magnitude of PAS in the device was the same as the one calculated for the components of the ECC tubing system, chosen as a term of comparison for their extensive diffusion. Interpolating the mean PAS values with respect to the number of passes, we obtained a straightforward prediction of the thrombogenic potential as a function of the duration of ECC.

  6. Effects of amount and type of dietary fats on postprandial lipemia and thrombogenic markers in individuals with metabolic syndrome.

    Science.gov (United States)

    Teng, Kim-Tiu; Chang, Chee-Yan; Kanthimathi, M S; Tan, Alexander Tong Boon; Nesaretnam, Kalanithi

    2015-09-01

    Postprandial lipemia has been reported to affect endothelial function by thrombogenic and inflammatory pathways. We set out to investigate the impact of a) specific amount (50 g vs 20 g fat), and b) type of fatty acids (saturated, monounsaturated or n-6 polyunsaturated fatty acids; SFA, MUFA, PUFA) on postprandial lipemia, thrombogenic and inflammatory factors in metabolic syndrome subjects. 30 subjects (15 men, 15 women) participated in a double-blind, randomized crossover design study with both the subjects and investigators blinded to treatments. Blood samples were collected at fasting and 30 min, hourly interval for a total of 6 h. As expected, lower triacylglycerol response was observed for low fat/high carbohydrate meal; whereas no difference was detected between the types of fatty acids. The incremental area under the curve (iAUC) for low fat/high carbohydrate meal was 70%, 81% and 61% lower than the SFA, MUFA and PUFA meals, respectively. The iAUC 0-6 h for triacylglycerol was 42% lower in women compared with the men (P = 0.024), with the similar trend observed for non-esterified fatty acids. There were significant meal × time interaction (P = 0.000) for plasma plasminogen activator inhibitor-1 and thromboxane B2 (P = 0.022) from baseline. No differences were observed between meals for plasma D-dimer, interleukin-6, interleukin-1β, tumor necrosis factor-α and high sensitivity C-reactive protein. These data indicate that in metabolic syndrome subjects, only the amount of dietary fatty acids affects postprandial lipemia but both amount and type of dietary fats alter thrombogenic factors. The study was registered at Clinicaltrials.gov (NCT01571947). Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  7. Aortic valve bypass

    DEFF Research Database (Denmark)

    Lund, Jens T; Jensen, Maiken Brit; Arendrup, Henrik

    2013-01-01

    In aortic valve bypass (AVB) a valve-containing conduit is connecting the apex of the left ventricle to the descending aorta. Candidates are patients with symptomatic aortic valve stenosis rejected for conventional aortic valve replacement (AVR) or transcatheter aortic valve implantation (TAVI). ...

  8. Liraglutide in polycystic ovary syndrome: a randomized trial, investigating effects on thrombogenic potential

    Directory of Open Access Journals (Sweden)

    Malin Nylander

    2017-01-01

    Full Text Available Polycystic ovary syndrome (PCOS is associated with increased risk of venous thromboembolism (VTE and cardiovascular disease (CVD in later life. We aimed to study the effect of liraglutide intervention on markers of VTE and CVD risk, in PCOS. In a double-blind, placebo-controlled, randomized trial, 72 overweight and/or insulinresistant women with PCOS were randomized, in a 2:1 ratio, to liraglutide or placebo 1.8 mg/day. Endpoints included between-group difference in change (baseline to follow-up in plasminogen activator inhibitor-1 levels and in thrombin generation test parameters: endogenous thrombin potential, peak thrombin concentration, lag time and time to peak. Mean weight loss was 5.2 kg (95% CI 3.0–7.5 kg, P < 0.001 in the liraglutide group compared with placebo. We detected no effect on endogenous thrombin potential in either group. In the liraglutide group, peak thrombin concentration decreased by 16.71 nmol/L (95% CI 2.32–31.11, P < 0.05 and lag time and time to peak increased by 0.13 min (95% CI 0.01–0.25, P < 0.05 and 0.38 min (95% CI 0.09–0.68, P < 0.05, respectively, but there were no between-group differences. There was a trend toward 12% (95% CI 0–23, P = 0.05 decreased plasminogen activator inhibitor-1 in the liraglutide group, and there was a trend toward 16% (95% CI −4 to 32, P = 0.10 reduction, compared with placebo. In overweight women with PCOS, liraglutide intervention caused an approximate 5% weight loss. In addition, liraglutide affected thrombin generation, although not significantly differently from placebo. A concomitant trend toward improved fibrinolysis indicates a possible reduction of the baseline thrombogenic potential. The findings point toward beneficial effects of liraglutide on markers of VTE and CVD risk, which should be further pursued in larger studies.

  9. Transcatheter aortic valve replacement

    Science.gov (United States)

    ... gov/ency/article/007684.htm Transcatheter aortic valve replacement To use the sharing features on this page, please enable JavaScript. Transcatheter aortic valve replacement (TAVR) is surgery to replace the aortic valve. ...

  10. Magnetically operated check valve

    Science.gov (United States)

    Morris, Brian G.; Bozeman, Richard J., Jr.

    1994-06-01

    A magnetically operated check valve is disclosed. The valve is comprised of a valve body and a movable poppet disposed therein. A magnet attracts the poppet to hold the valve shut until the force of fluid flow through the valve overcomes the magnetic attraction and moves the poppet to an unseated, open position. The poppet and magnet are configured and disposed to trap a magnetically attracted particulate and prevent it from flowing to a valve seating region.

  11. [Tricuspid valve insufficiency: what should be done?].

    Science.gov (United States)

    von Segesser, L K; Stauffer, J C; Delabays, A; Chassot, P G

    1998-12-01

    Tricuspid regurgitation is relatively common. Due to the progress made in echocardiography, its diagnosis is in general made readily and in reliable fashion. Basically one has to distinguish between functional tricuspid valve regurgitation due to volume and/or pressure overload of the right ventricle with intact valve structures versus tricuspid valve regurgitation due to pathologic valve structures. The clear identification of the regurgitation mechanism is of prime importance for the treatment. Functional tricuspid valve regurgitation can often be improved by medical treatment of heart failure, and eventually a tricuspid valve plasty can solve the problem. However, the presence of pathologic tricuspid valve structures makes in general more specific plastic surgical procedures and even prosthetic valve replacements necessary. A typical example for a structural tricuspid valve regurgitation is the case of a traumatic papillary muscle rupture. Due to the sudden onset, this pathology is not well tolerated and requires in general surgical reinsertion of the papillary muscle. In contrast, tricuspid valve regurgitation resulting from chronic pulmonary embolism with pulmonary artery hypertension, can be improved by pulmonary artery thrombendarteriectomy and even completely cured with an additional tricuspid annuloplasty. However, tricuspid regurgitations due to terminal heart failure are not be addressed with surgery directed to tricuspid valve repair or replacement. Heart transplantation, dynamic cardiomyoplasty or mechanical circulatory support should be evaluated instead.

  12. What Is Heart Valve Surgery?

    Science.gov (United States)

    ... working correctly. Most valve replacements involve the aortic Tricuspid valve and mitral valves. The aortic valve separates ... where it shouldn’t. This is called incompetence, insufficiency or regurgitation. • Prolapse — mitral valve flaps don’t ...

  13. What Is Heart Valve Disease?

    Science.gov (United States)

    ... and replacing it with a man-made or biological valve. Biological valves are made from pig, cow, or human ... the valve. Man-made valves last longer than biological valves and usually don’t have to be ...

  14. Water hammer caused by closure of turbine safety spherical valves

    Science.gov (United States)

    Karadžić, U.; Bergant, A.; Vukoslavčević, P.

    2010-08-01

    This paper investigates water hammer effects caused by closure of spherical valves against the discharge. During the first phase of modernisation of Perućica high-head hydropower plant (HPP), Montenegro, safety spherical valves (inlet turbine valves) have been refurbished on the first two Pelton turbine units. The valve closure is controlled by the valve actuator (hydraulic servomotor). Because the torque acting on the valve body is dependent on flow conditions the valve closing time may vary significantly for different flow velocities (passive valve). For the passive valve the torques acting on the valve body should be considered in the valve model. The valve closing time results from numerical simulation. On the contrary, for the active valve the valve closing time is assumed prior to simulation. The spherical valve boundary condition is incorporated into the method of characteristics (MOC) algorithm. The staggered (diamond) grid in applying the MOC is used in this paper. The passive valve boundary condition is described by the water hammer equations, the valve equation that relates discharge to pressure head drop and the dynamic equation of the valve body motion (torque equation). The active valve boundary condition is described by the first two equations, respectively. Standard quasi-steady friction model is used for estimating friction losses in plant's tunnel and penstocks. Numerical results using both the active and the passive spherical valve models are compared with results of measurements. It has been found that the influence of flow conditions on the spherical valve closing time is minor for the cases considered. Computed and measured results agree reasonably well.

  15. Water hammer caused by closure of turbine safety spherical valves

    International Nuclear Information System (INIS)

    Karadzic, U; Vukoslavcevic, P; Bergant, A

    2010-01-01

    This paper investigates water hammer effects caused by closure of spherical valves against the discharge. During the first phase of modernisation of Perucica high-head hydropower plant (HPP), Montenegro, safety spherical valves (inlet turbine valves) have been refurbished on the first two Pelton turbine units. The valve closure is controlled by the valve actuator (hydraulic servomotor). Because the torque acting on the valve body is dependent on flow conditions the valve closing time may vary significantly for different flow velocities (passive valve). For the passive valve the torques acting on the valve body should be considered in the valve model. The valve closing time results from numerical simulation. On the contrary, for the active valve the valve closing time is assumed prior to simulation. The spherical valve boundary condition is incorporated into the method of characteristics (MOC) algorithm. The staggered (diamond) grid in applying the MOC is used in this paper. The passive valve boundary condition is described by the water hammer equations, the valve equation that relates discharge to pressure head drop and the dynamic equation of the valve body motion (torque equation). The active valve boundary condition is described by the first two equations, respectively. Standard quasi-steady friction model is used for estimating friction losses in plant's tunnel and penstocks. Numerical results using both the active and the passive spherical valve models are compared with results of measurements. It has been found that the influence of flow conditions on the spherical valve closing time is minor for the cases considered. Computed and measured results agree reasonably well.

  16. Water hammer caused by closure of turbine safety spherical valves

    Energy Technology Data Exchange (ETDEWEB)

    Karadzic, U; Vukoslavcevic, P [Faculty of Mechanical Engineering, University of Montenegro Dzordza Vasingtona nn, Podgorica, 81000 (Montenegro); Bergant, A, E-mail: uros.karadzic@ac.m [LitostrojPower d.o.o., Litostrojska 50, Ljubljana, 1000 (Slovenia)

    2010-08-15

    This paper investigates water hammer effects caused by closure of spherical valves against the discharge. During the first phase of modernisation of Perucica high-head hydropower plant (HPP), Montenegro, safety spherical valves (inlet turbine valves) have been refurbished on the first two Pelton turbine units. The valve closure is controlled by the valve actuator (hydraulic servomotor). Because the torque acting on the valve body is dependent on flow conditions the valve closing time may vary significantly for different flow velocities (passive valve). For the passive valve the torques acting on the valve body should be considered in the valve model. The valve closing time results from numerical simulation. On the contrary, for the active valve the valve closing time is assumed prior to simulation. The spherical valve boundary condition is incorporated into the method of characteristics (MOC) algorithm. The staggered (diamond) grid in applying the MOC is used in this paper. The passive valve boundary condition is described by the water hammer equations, the valve equation that relates discharge to pressure head drop and the dynamic equation of the valve body motion (torque equation). The active valve boundary condition is described by the first two equations, respectively. Standard quasi-steady friction model is used for estimating friction losses in plant's tunnel and penstocks. Numerical results using both the active and the passive spherical valve models are compared with results of measurements. It has been found that the influence of flow conditions on the spherical valve closing time is minor for the cases considered. Computed and measured results agree reasonably well.

  17. Tricuspid valve endocarditis

    Science.gov (United States)

    Hussain, Syed T.; Witten, James; Shrestha, Nabin K.; Blackstone, Eugene H.

    2017-01-01

    Right-sided infective endocarditis (RSIE) is less common than left-sided infective endocarditis (IE), encompassing only 5–10% of cases of IE. Ninety percent of RSIE involves the tricuspid valve (TV). Given the relatively small numbers of TVIE cases operated on at most institutions, the purpose of this review is to highlight and discuss the current understanding of IE involving the TV. RSIE and TVIE are strongly associated with intravenous drug use (IVDU), although pacemaker leads, defibrillator leads and vascular access for dialysis are also major risk factors. Staphylococcus aureus is the predominant causative organism in TVIE. Most patients with TVIE are successfully treated with antibiotics, however, 5–16% of RSIE cases eventually require surgical intervention. Indications and timing for surgery are less clear than for left-sided IE; surgery is primarily considered for failed medical therapy, large vegetations and septic pulmonary embolism, and less often for TV regurgitation and heart failure. Most patients with an infected prosthetic TV will require surgery. Concomitant left-sided IE has its own surgical indications. Earlier surgical intervention may potentially prevent further destruction of leaflet tissue and increase the likelihood of TV repair. Fortunately, TV debridement and repair can be accomplished in most cases, even those with extensive valve destruction, using a variety of techniques. Valve repair is advocated over replacement, particularly in IVDUs patients who are young, non-compliant and have a higher risk of recurrent infection and reoperation with valve replacement. Excising the valve without replacing, it is not advocated; it has been reported previously, but these patients are likely to be symptomatic, particularly in cases with septic pulmonary embolism and increased pulmonary vascular resistance. Patients with concomitant left-sided involvement have worse prognosis than those with RSIE alone, due predominantly to greater likelihood of

  18. Procedural Experience for Transcatheter Aortic Valve Replacement and Relation to Outcomes: The STS/ACC TVT Registry.

    Science.gov (United States)

    Carroll, John D; Vemulapalli, Sreekanth; Dai, Dadi; Matsouaka, Roland; Blackstone, Eugene; Edwards, Fred; Masoudi, Frederick A; Mack, Michael; Peterson, Eric D; Holmes, David; Rumsfeld, John S; Tuzcu, E Murat; Grover, Frederick

    2017-07-04

    Transcatheter aortic valve replacement (TAVR) has been introduced into U.S. clinical practice with efforts to optimize outcomes and minimize the learning curve. The goal of this study was to assess the degree to which increasing experience during the introduction of this procedure, separated from other outcome determinants including patient and procedural characteristics, is associated with outcomes. The authors evaluated the association of hospital TAVR volume and patient outcomes for TAVR by using data from 42,988 commercial procedures conducted at 395 hospitals submitting to the Transcatheter Valve Therapy Registry from 2011 through 2015. Outcomes assessed included adjusted and unadjusted in-hospital major adverse events. Increasing site volume was associated with lower in-hospital risk-adjusted outcomes, including mortality (p < 0.02), vascular complications (p < 0.003), and bleeding (p < 0.001) but was not associated with stroke (p = 0.14). From the first case to the 400th case in the volume-outcome model, risk-adjusted adverse outcomes declined, including mortality (3.57% to 2.15%), bleeding (9.56% to 5.08%), vascular complications (6.11% to 4.20%), and stroke (2.03% to 1.66%). Vascular and bleeding volume-outcome associations were nonlinear with a higher risk of adverse outcomes in the first 100 cases. An association of procedure volume with risk-adjusted outcomes was also seen in the subgroup having transfemoral access. The initial adoption of TAVR into practice in the United States showed that increasing experience was associated with better outcomes. This association, whether deemed a prolonged learning curve or a manifestation of a volume-outcome relationship, suggested that concentrating experience in higher volume heart valve centers might be a means of improving outcomes. (STS/ACC Transcatheter Valve Therapy Registry [TVT Registry]; NCT01737528). Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All

  19. Microfluidic sieve valves

    Science.gov (United States)

    Quake, Stephen R; Marcus, Joshua S; Hansen, Carl L

    2015-01-13

    Sieve valves for use in microfluidic device are provided. The valves are useful for impeding the flow of particles, such as chromatography beads or cells, in a microfluidic channel while allowing liquid solution to pass through the valve. The valves find particular use in making microfluidic chromatography modules.

  20. Rotary pneumatic valve

    Science.gov (United States)

    Hardee, Harry C.

    1991-01-01

    A rotary pneumatic valve which is thrust balanced and the pneumatic pressure developed produces only radial loads on the valve cylinder producing negligible resistance and thus minimal torque on the bearings of the valve. The valve is multiplexed such that at least two complete switching cycles occur for each revolution of the cylinder spindle.

  1. Mitral Valve Stenosis

    Science.gov (United States)

    ... the left ventricle from flowing backward. A defective heart valve fails to either open or close fully. Risk factors Mitral valve stenosis is less common today than it once was because the most common cause, ... other heart valve problems, mitral valve stenosis can strain your ...

  2. Aortic Valve Stenosis

    Science.gov (United States)

    ... most cases, doctors don't know why a heart valve fails to develop properly, so it isn't something you could have prevented. Calcium buildup on the valve. With age, heart valves may accumulate deposits of calcium (aortic valve ...

  3. Remote actuated valve implant

    Science.gov (United States)

    McKnight, Timothy E; Johnson, Anthony; Moise, Jr., Kenneth J; Ericson, Milton Nance; Baba, Justin S; Wilgen, John B; Evans, III, Boyd McCutchen

    2014-02-25

    Valve implant systems positionable within a flow passage, the systems having an inlet, an outlet, and a remotely activatable valve between the inlet and outlet, with the valves being operable to provide intermittent occlusion of the flow path. A remote field is applied to provide thermal or magnetic activation of the valves.

  4. Induced pluripotent stem cells with NOTCH1 gene mutation show impaired differentiation into smooth muscle and endothelial cells: Implications for bicuspid aortic valve-related aortopathy.

    Science.gov (United States)

    Jiao, Jiao; Tian, Weihua; Qiu, Ping; Norton, Elizabeth L; Wang, Michael M; Chen, Y Eugene; Yang, Bo

    2018-03-12

    The NOTCH1 gene mutation has been identified in bicuspid aortic valve patients. We developed an in vitro model with human induced pluripotent stem cells (iPSCs) to evaluate the role of NOTCH1 in smooth muscle and endothelial cell (EC) differentiation. The iPSCs were derived from a patient with a normal tricuspid aortic valve and aorta. The NOTCH1 gene was targeted in iPSCs with the Clustered Regularly Interspaced Short Palindromic Repeats/CRISPR-associated protein 9 nuclease (Cas9) system. The NOTCH1 -/- (NOTCH1 homozygous knockout) and isogenic control iPSCs (wild type) were differentiated into neural crest stem cells (NCSCs) and into cardiovascular progenitor cells (CVPCs). The NCSCs were differentiated into smooth muscle cells (SMCs). The CVPCs were differentiated into ECs. The differentiations of SMCs and ECs were compared between NOTCH1 -/- and wild type cells. The expression of NCSC markers (SRY-related HMG-box 10 and transcription factor AP-2 alpha) was significantly lower in NOTCH1 -/- NCSCs than in wild type NCSCs. The SMCs derived from NOTCH1 -/- NCSCs showed immature morphology with smaller size and decreased expression of all SMC-specific contractile proteins. In NOTCH1 -/- CVPCs, the expression of ISL1, NKX2.5, and MYOCD was significantly lower than that in isogenic control CVPCs, indicating impaired differentiation from iPSCs to CVPCs. The NOTCH1 -/- ECs derived from CVPCs showed significantly lower expression of cluster of differentiation 105 and cluster of differentiation 31 mRNA and protein, indicating a defective differentiation process. NOTCH1 is critical in SMC and EC differentiation of iPSCs through NCSCs and CVPCs, respectively. NOTCH1 gene mutations might potentially contribute to the development of thoracic aortic aneurysms by affecting SMC differentiation in some patients with bicuspid aortic valve. Copyright © 2018 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

  5. Scissor thrust valve actuator

    Science.gov (United States)

    DeWall, Kevin G.; Watkins, John C; Nitzel, Michael E.

    2006-08-29

    Apparatus for actuating a valve includes a support frame and at least one valve driving linkage arm, one end of which is rotatably connected to a valve stem of the valve and the other end of which is rotatably connected to a screw block. A motor connected to the frame is operatively connected to a motor driven shaft which is in threaded screw driving relationship with the screw block. The motor rotates the motor driven shaft which drives translational movement of the screw block which drives rotatable movement of the valve driving linkage arm which drives translational movement of the valve stem. The valve actuator may further include a sensory control element disposed in operative relationship with the valve stem, the sensory control element being adapted to provide control over the position of the valve stem by at least sensing the travel and/or position of the valve stem.

  6. Corrosion, wettability and thrombogenicity investigation of ion beam modified HAP/Al{sub 2}O{sub 3}

    Energy Technology Data Exchange (ETDEWEB)

    Shikha, Deep; Shahid, Md.; Jha, Ush [Department of Chemistry, BIT Mesra, Ranchi 835215 (India); Sinha, Sanjay Kumar, E-mail: sksinha@bitmesra.ac.in [Department of Physics, BIT Mesra, Ranchi 835215 (India); Reddy, V. Raghavendra [UGC-DAE Consortium for Scientific Research Indore (India); Ojha, Sunil; Kumar, P.; Kanjilal, Dinakar [IUAC, Aruna Asaf Ali road, New Delhi (India)

    2015-08-01

    Hydroxyapatite Ca{sub 10}(PO4){sub 6}(OH){sub 2}, is a biosensitive ceramic which promotes bone growth in human fluid. However due to its weak mechanical strength it is often coated on chemically inert material such as alumina. In the present work hydroxyapatite (HAP) of thickness 1 μm is deposited on alumina using sol–gel process and the stoichiometric ratio of Ca:P = 1.67 is maintained. Surface property of HAP is further improved by Ar{sup +} ion implantation. The samples were irradiated with 1.4 MeV Ar{sup +} ions with fluence ranging from 5 × 10{sup 14} to 1 × 10{sup 16} ions/cm{sup 2}. After irradiation, the surface is characterized using Atomic Force Microscope (AFM), Scanning Electron Microscope (SEM), electron dispersive X-ray spectroscopy (EDX), Glancing incidence X-ray diffraction (GXRD) and Rutherford backscattering Spectroscopy (RBS) techniques. Corrosion resistances and impedance analysis were carried out in Ringer solution. RBS and EDX were used to confirm the stoichiometric ratio of the film. Compound formation before and after ion implantation was studied using GXRD. Nanohardness, wettability and thrombogenicity of all the samples were studied. Correlation among surface morphology, improvement in corrosion resistance, hardness, wetability and thrombogenicity before and after ion implantation are discussed in this paper. - Highlights: • .People have worked on HAP coated on metallic alloy and even alumina but the characterization done here are all different. • Throbmobogenicity, corrosion resistance and wetability have all been carried out first time. • Improvement of Surface and interface using energetic inert ions like argon is carried out first time. • The best ion fluence for orthopaedic implants is proposed.

  7. The radiology of prosthetic heart valves

    International Nuclear Information System (INIS)

    Steiner, R.M.; Flicker, S.

    1985-01-01

    The development of prosthetic heart valves in the late 1950s ushered in a new era in the treatment of heart disease. The radiologist has an important role to play preoperatively in the diagnosis of valvular heart disease. Radiology is valuable in identification of the implanted prosthetic valve and recognition of complications associated with valve implantation. Radiologists must be familiar with the imaging techniques best suited to evaluate the function of the valve prosthesis in question. In this chapter the authors discuss the radiographic approach to the evaluation of the status of patients for valve replacement and the imaging problems peculiar to the types of valves in current use. The relative value of plain-film radiography, fluoroscopy, videorecording and cinerecording, and aortography is addressed, as well as the potential value of magnetic resonance imaging and subsecond dynamic computed tomography

  8. PNGS-A unit 4 composite graphite valve packing program

    International Nuclear Information System (INIS)

    Meaney, F.

    1992-01-01

    During the outage of Pickering 4 for retubing and rehabilitation, a composite graphite packing program was completed on 100 conventional valves and 50 nuclear valves. This paper describes component refurbishment, related maintenance, and manpower requirements. 2 figs

  9. Bioprosthetic Valve Fracture During Valve-in-valve TAVR: Bench to Bedside.

    Science.gov (United States)

    Saxon, John T; Allen, Keith B; Cohen, David J; Chhatriwalla, Adnan K

    2018-01-01

    Valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) has been established as a safe and effective means of treating failed surgical bioprosthetic valves (BPVs) in patients at high risk for complications related to reoperation. Patients who undergo VIV TAVR are at risk of patient-prosthesis mismatch, as the transcatheter heart valve (THV) is implanted within the ring of the existing BPV, limiting full expansion and reducing the maximum achievable effective orifice area of the THV. Importantly, patient-prosthesis mismatch and high residual transvalvular gradients are associated with reduced survival following VIV TAVR. Bioprosthetic valve fracture (BVF) is as a novel technique to address this problem. During BPV, a non-compliant valvuloplasty balloon is positioned within the BPV frame, and a highpressure balloon inflation is performed to fracture the surgical sewing ring of the BPV. This allows for further expansion of the BPV as well as the implanted THV, thus increasing the maximum effective orifice area that can be achieved after VIV TAVR. This review focuses on the current evidence base for BVF to facilitate VIV TAVR, including initial bench testing, procedural technique, clinical experience and future directions.

  10. Double-reed exhaust valve engine

    Science.gov (United States)

    Bennett, Charles L.

    2015-06-30

    An engine based on a reciprocating piston engine that extracts work from pressurized working fluid. The engine includes a double reed outlet valve for controlling the flow of low-pressure working fluid out of the engine. The double reed provides a stronger force resisting closure of the outlet valve than the force tending to open the outlet valve. The double reed valve enables engine operation at relatively higher torque and lower efficiency at low speed, with lower torque, but higher efficiency at high speed.

  11. Which valve is which?

    Directory of Open Access Journals (Sweden)

    Pravin Saxena

    2015-01-01

    Full Text Available A 25-year-old man presented with a history of breathlessness for the past 2 years. He had a history of operation for Tetralogy of Fallot at the age of 5 years and history suggestive of Rheumatic fever at the age of 7 years. On echocardiographic examination, all his heart valves were severely regurgitating. Morphologically, all the valves were irreparable. The ejection fraction was 35%. He underwent quadruple valve replacement. The aortic and mitral valves were replaced by metallic valve and the tricuspid and pulmonary by tissue valve.

  12. Bioprosthetic Valve Fracture Improves the Hemodynamic Results of Valve-in-Valve Transcatheter Aortic Valve Replacement.

    Science.gov (United States)

    Chhatriwalla, Adnan K; Allen, Keith B; Saxon, John T; Cohen, David J; Aggarwal, Sanjeev; Hart, Anthony J; Baron, Suzanne J; Dvir, Danny; Borkon, A Michael

    2017-07-01

    Valve-in-valve (VIV) transcatheter aortic valve replacement (TAVR) may be less effective in small surgical valves because of patient/prosthesis mismatch. Bioprosthetic valve fracture (BVF) using a high-pressure balloon can be performed to facilitate VIV TAVR. We report data from 20 consecutive clinical cases in which BVF was successfully performed before or after VIV TAVR by inflation of a high-pressure balloon positioned across the valve ring during rapid ventricular pacing. Hemodynamic measurements and calculation of the valve effective orifice area were performed at baseline, immediately after VIV TAVR, and after BVF. BVF was successfully performed in 20 patients undergoing VIV TAVR with balloon-expandable (n=8) or self-expanding (n=12) transcatheter valves in Mitroflow, Carpentier-Edwards Perimount, Magna and Magna Ease, Biocor Epic and Biocor Epic Supra, and Mosaic surgical valves. Successful fracture was noted fluoroscopically when the waist of the balloon released and by a sudden drop in inflation pressure, often accompanied by an audible snap. BVF resulted in a reduction in the mean transvalvular gradient (from 20.5±7.4 to 6.7±3.7 mm Hg, P valve effective orifice area (from 1.0±0.4 to 1.8±0.6 cm 2 , P valves to facilitate VIV TAVR with either balloon-expandable or self-expanding transcatheter valves and results in reduced residual transvalvular gradients and increased valve effective orifice area. © 2017 American Heart Association, Inc.

  13. Experimental Assessment of Flow Fields Associated with Heart Valve Prostheses Using Particle Image Velocimetry (PIV): Recommendations for Best Practices.

    Science.gov (United States)

    Raghav, Vrishank; Sastry, Sudeep; Saikrishnan, Neelakantan

    2018-03-12

    Experimental flow field characterization is a critical component of the assessment of the hemolytic and thrombogenic potential of heart valve substitutes, thus it is important to identify best practices for these experimental techniques. This paper presents a brief review of commonly used flow assessment techniques such as Particle image velocimetry (PIV), Laser doppler velocimetry, and Phase contrast magnetic resonance imaging and a comparison of these methodologies. In particular, recommendations for setting up planar PIV experiments such as recommended imaging instrumentation, acquisition and data processing are discussed in the context of heart valve flows. Multiple metrics such as residence time, local velocity and shear stress that have been identified in the literature as being relevant to hemolysis and thrombosis in heart valves are discussed. Additionally, a framework for uncertainty analysis and data reporting for PIV studies of heart valves is presented in this paper. It is anticipated that this paper will provide useful information for heart valve device manufacturers and researchers to assess heart valve flow fields for the potential for hemolysis and thrombosis.

  14. AREVA's innovative solutions for valve diagnostics and in-situ valve repair

    International Nuclear Information System (INIS)

    Damies, H.; Breitenberger, U.; Munoz, L.; Kostroun, F.

    2012-01-01

    Optimized maintenance strategies are a key aspect for safe and undisturbed plant operation. Innovative valve service solutions can support that in an efficient way. The ADAM®/SIPLUG® valve monitoring system allows full online monitoring of valves and actuators with automatic evaluation and assessment. Especially for safety-related and operation-related valves this provides valuable information on components condition to ensure proper function and contribute to optimization of maintenance strategies as well as effective maintenance performance. More than 25 years of experience in various plants worldwide show that application of ADAM®/SIPLUG® valve diagnostics solution leads to increased plant safety and availability. With the innovative AVARIS technology an in-situ valve repair is possible. It has the unique ability to conduct several steps in-situ, to maintain the sealing seat of gate or check valves. By applying AVARIS, the valve is restored in its original state, the system remains unchanged. Thus, all original documents remain valid and applicable. In comparison to previous procedures like cutting valves out of the pipeline and repairing hard facings or damaged seal seats in a separate workshop or alternatively replacement by a new valve body the new AVARIS technology avoids costs, risk and effort. (author)

  15. Sealing a Loosely Fitting Valve Assembly

    Science.gov (United States)

    Goff, L.; Tellier, G.

    1986-01-01

    Double-ring seal avoids expense of remachining or redesigning valve parts. Mating fittings on valve sealed by pair of rings - one O-ring and backup ring. Backup ring fills relatively large gap between parts. Prevents softer O-ring from being pushed into and through gap.

  16. Mitral Valve Prolapse

    Science.gov (United States)

    ... valve syndrome . What happens during MVP? Watch an animation of mitral valve prolapse When the heart pumps ( ... our brochures Popular Articles 1 Understanding Blood Pressure Readings 2 Sodium and Salt 3 Heart Attack Symptoms ...

  17. Problem: Mitral Valve Regurgitation

    Science.gov (United States)

    ... each time the left ventricle contracts. Watch an animation of mitral valve regurgitation A leaking mitral valve ... Not Alone Popular Articles 1 Understanding Blood Pressure Readings 2 Sodium and Salt 3 Heart Attack Symptoms ...

  18. Problem: Heart Valve Regurgitation

    Science.gov (United States)

    ... should be completely closed For example: Watch an animation of mitral valve regurgitation A leaking mitral valve ... Not Alone Popular Articles 1 Understanding Blood Pressure Readings 2 Sodium and Salt 3 Heart Attack Symptoms ...

  19. Aortic valve surgery - open

    Science.gov (United States)

    ... gov/ency/article/007408.htm Aortic valve surgery - open To use the sharing features on this page, ... separates the heart and aorta. The aortic valve opens so blood can flow out. It then closes ...

  20. Mitral valve surgery - open

    Science.gov (United States)

    ... Taking warfarin (Coumadin) References Otto CM, Bonow RO. Valvular heart disease. In: Mann DL, Zipes DP, Libby P, Bonow ... A.M. Editorial team. Heart Surgery Read more Heart Valve Diseases Read more Mitral Valve Prolapse Read more A. ...

  1. Swing check valve

    International Nuclear Information System (INIS)

    Eminger, H.E.

    1977-01-01

    A swing check valve which includes a valve body having an inlet and outlet is described. A recess in the valve body designed to hold a seal ring and a check valve disc swingable between open and closed positions. The disc is supported by a high strength wire secured at one end in a support spacer pinned through bearing blocks fixed to the valve body and at its other end in a groove formed on the outer peripheral surface of the disc. The parts are designed and chosen such to provide a lightweight valve disc which is held open by minimum velocity of fluid flowing through the valve which thus reduces oscillations and accompanying wear of bearings supporting the valve operating parts. (Auth.)

  2. Relation of Serum Vitamin D to Risk of Mitral Annular and Aortic Valve Calcium (from the Multi-Ethnic Study of Atherosclerosis).

    Science.gov (United States)

    Tibuakuu, Martin; Zhao, Di; de Boer, Ian H; Guallar, Eliseo; Bortnick, Anna E; Lutsey, Pamela L; Budoff, Matthew J; Kizer, Jorge R; Kestenbaum, Bryan R; Michos, Erin D

    2017-08-01

    Serum 25-hydroxyvitamin D [25(OH)D] concentration has been identified as a possible modifiable risk factor for cardiovascular disease (CVD). We hypothesized that serum 25(OH)D concentration would be associated with calcifications of the left-sided heart valves, which are markers of CVD risk. Aortic valve calcium (AVC) and mitral annular calcium (MAC) were quantified from cardiac computed tomography scans performed on 5,530 Multi-Ethnic Study of Atherosclerosis participants at the baseline examination (2000 to 2002) and at a follow-up visit at either Examination 2 (2002 to 2004) or Examination 3 (2004 to 2005). 25(OH)D was measured from serum samples collected at the baseline examination. Using relative risk regression, we evaluated the multivariable-adjusted risk of prevalent and incident AVC and MAC in this ethnically diverse population free of clinical CVD at baseline. The mean age of participants was 62 ± 10 years; 53% were women, 40% white, 26% black, 21% Hispanic, and 12% Chinese. Prevalent AVC and MAC were observed in 12% and 9% of study sample, respectively. There were no significant associations between 25(OH)D and prevalent AVC or MAC. Over a mean follow-up of 2.5 years, 4% developed incident AVC and 5% developed incident MAC. After adjusting for demographic variables, each 10 ng/ml higher serum 25(OH)D was associated with a 15% (relative risk 0.85, 95% confidence interval 0.74 to 0.98) lower risk of incident MAC but not AVC. However, this association was no longer significant after adjusting for lifestyle and CVD risk factors. Results suggest a possible link between serum 25(OH)D and the risk for incident MAC, but future studies with longer follow-up are needed to further test this association. Copyright © 2017 Elsevier Inc. All rights reserved.

  3. A new methodological approach to assess cardiac work by pressure-volume and stress-length relations in patients with aortic valve stenosis and dilated cardiomyopathy.

    Science.gov (United States)

    Alter, P; Rupp, H; Rominger, M B; Klose, K J; Maisch, B

    2008-01-01

    In experimental animals, cardiac work is derived from pressure-volume area and analyzed further using stress-length relations. Lack of methods for determining accurately myocardial mass has until now prevented the use of stress-length relations in patients. We hypothesized, therefore, that not only pressure-volume loops but also stress-length diagrams can be derived from cardiac volume and cardiac mass as assessed by cardiac magnetic resonance imaging (CMR) and invasively measured pressure. Left ventricular (LV) volume and myocardial mass were assessed in seven patients with aortic valve stenosis (AS), eight with dilated cardiomyopathy (DCM), and eight controls using electrocardiogram (ECG)-gated CMR. LV pressure was measured invasively. Pressure-volume curves were calculated based on ECG triggering. Stroke work was assessed as area within the pressure-volume loop. LV wall stress was calculated using a thick-wall sphere model. Similarly, stress-length loops were calculated to quantify stress-length-based work. Taking the LV geometry into account, the normalization with regard to ventricular circumference resulted in "myocardial work." Patients with AS (valve area 0.73+/-0.18 cm(2)) exhibited an increased LV myocardial mass when compared with controls (Pwork of AS was unchanged when compared with controls (0.539+/-0.272 vs 0.621+/-0.138 Nm, not significant), whereas DCM exhibited a significant depression (0.367+/-0.157 Nm, Pwork was significantly reduced in both AS and DCM when compared with controls (129.8+/-69.6, 200.6+/-80.1, 332.2+/-89.6 Nm/m(2), Pmethodological approach of using CMR and invasive pressure measurement. Myocardial work was reduced in patients with DCM and noteworthy also in AS, while stroke work was reduced in DCM only. Most likely, deterioration of myocardial work is crucial for the prognosis. It is suggested to include these basic physiological procedures in the clinical assessment of the pump function of the heart.

  4. Mitral Valve Prolapse

    Science.gov (United States)

    Mitral valve prolapse (MVP) occurs when one of your heart's valves doesn't work properly. The flaps of the valve are "floppy" and ... to run in families. Most of the time, MVP doesn't cause any problems. Rarely, blood can ...

  5. Overflow control valve

    International Nuclear Information System (INIS)

    Kessinger, B.A.; Hundal, R.; Parlak, E.A.

    1982-01-01

    An overflow control valve for use in a liquid sodium coolant pump tank which can be remotely engaged with and disengaged from the pump tank wall to thereby permit valve removal. An actuating shaft for controlling the valve also has means for operating a sliding cylinder against a spring to retract the cylinder from sealing contact with the pump tank nozzle. (author)

  6. Fluid control valves

    International Nuclear Information System (INIS)

    Rankin, J.

    1980-01-01

    A fluid control valve is described in which it is not necessary to insert a hand or a tool into the housing to remove the valve seat. Such a valve is particularly suitable for the control of radioactive fluids since maintenance by remote control is possible. (UK)

  7. A remote control valve

    International Nuclear Information System (INIS)

    Cachard, Maurice de; Dumont, Maurice.

    1976-01-01

    This invention concerns a remote control valve for shutting off or distributing a fluid flowing at a high rate and low pressure. Among the different valves at present in use, electric valves are the most recommended for remote control but their reliability is uncertain and they soon become costly when large diameter valves are used. The valve described in this invention does away with this drawback owing to its simplicity and the small number of moving parts, this makes it particularly reliable. It mainly includes: a tubular body fitted with at least one side opening; at least one valve wedge for this opening, coaxial with the body, and mobile; a mobile piston integral with this wedge. Several valves to the specifications of this invention can be fitted in series (a shut-off valve can be used in conjunction with one or more distribution valves). The fitting and maintenance of the valve is very simple owing to its design. It can be fabricated in any material such as metals, alloys, plastics and concrete. The structure of the valve prevents the flowing fluid from coming into contact with the outside environment, thereby making it particularly suitable in the handling of dangerous or corrosive fluids. Finally, the opening and shutting of the valve occurs slowly, thereby doing away with the water hammer effect so frequent in large bore pipes [fr

  8. Heart Valve Diseases

    Science.gov (United States)

    Your heart has four valves. Normally, these valves open to let blood flow through or out of your heart, and then shut to keep it from flowing ... close tightly. It's one of the most common heart valve conditions. Sometimes it causes regurgitation. Stenosis - when ...

  9. Plunger with simple retention valve

    International Nuclear Information System (INIS)

    Fekete, A.V.

    1987-01-01

    This patent describes a positive displacement retention valve apparatus in which the actual flow equals the theoretical maximum flow through the retention valve. The apparatus includes, in combination, a confined fluid flow conduit, a piston adapted for reciprocal movement within the fluid flow conduit between upstream and downstream limit positions, piston reciprocating means, and pressure responsive check valve means located upstream with respect to the piston in the fluid flow conduit. The pressure responsive check valve means operable to permit fluid flow therethrough in a downstream direction toward the piston, and to preclude fluid flow therethrough in an opposite direction. The piston is composed of parts which are relatively movable with respect to one another. The piston includes a simple retention valve consisting of a plug means, a cylinder having a minimum and a maximum internal cross section flow area therein and being reciprocal within the confined fluid flow conduit, and a seat on the cylinder for the plug means. The piston reciprocating means are operatively connected to the plug means

  10. Fatty acid and cholesterol profiles and hypocholesterolemic, atherogenic, and thrombogenic indices of table eggs in the retail market.

    Science.gov (United States)

    Attia, Youssef A; Al-Harthi, Mohammed A; Korish, Mohamed A; Shiboob, Mohamed M

    2015-10-27

    Eggs are an important source of food due to its favorable effects on human health derived from the protein, fats, minerals, vitamins and bioactive components. We studied the effects of source of eggs in the retail market on fatty acids, lipid profiles and antioxidant status in eggs. Eggs from four sources named A, B, C, and D in the retail market were collected to determine fatty acid, total lipid, and cholesterol profiles; hypocholesterolemic, atherogenic and thrombotic indices; antioxidant status (e.g., of malondialdehyde); and total antioxidant capacity in the whole edible parts of eggs (albumen + yolk) and egg yolk. Samples were collected four times and pooled over times to represent 5 and 10 samples per source for determinations of fatty acids and determinations of lipid profiles and antioxidant status, respectively. Fatty acid, total lipid, and cholesterol profiles; hypocholesterolemic, atherogenic and thrombotic indices; presence of malondialdehyde; and total antioxidant capacity in the whole edible parts of eggs and egg yolk showed significant differences (P ≥ 0.05) among different sources of eggs in retail market. Source D showed higher levels of saturated fatty acids (SFA) and linoleic and monounsaturated fatty acid (MUFA)/polyunsaturated fatty acid (PUFA) ratio but lower levels of MUFA and linolenic, arachidonic, eicosapentaeonic (EPA), decohexaenoic (DHA), and total ω9 fatty acids and lower unsaturated fatty acids (UFA)/SFA ratio. Similar trend was shown in fatty acids profiles of the whole edible parts of eggs. On the other hand, total cholesterol, low density lipoprotein (LDL), LDL/high density lipoprotein (HDL) ratio, and atherogenic and thrombogenic indices and total antioxidant capacity of source D were significantly higher than those of other source, but levels of hypocholesterolemic index, and malondialdehyde levels were lower for source D. Eggs in the retail market in Jeddah city, Saudi Arabia, from May to August 2015 showed a

  11. Characterization of a bioprosthetic bicuspid venous valve hemodynamics: implications for mechanism of valve dynamics.

    Science.gov (United States)

    Tien, W-H; Chen, H Y; Berwick, Z C; Krieger, J; Chambers, S; Dabiri, D; Kassab, G S

    2014-10-01

    Chronic venous insufficiency (CVI) of the lower extremities is a common clinical problem. Although bioprosthetic valves have been proposed to treat severe reflux, clinical success has been limited due to thrombosis and neointima overgrowth of the leaflets that is, in part, related to the hemodynamics of the valve. A bioprosthetic valve that mimics native valve hemodynamics is essential. A computational model of the prosthetic valve based on realistic geometry and mechanical properties was developed to simulate the interaction of valve structure (fluid-structure interaction, FSI) with the surrounding flow. The simulation results were validated by experiments of a bioprosthetic bicuspid venous valve using particle image velocimetry (PIV) with high spatial and temporal resolution in a pulse duplicator (PD). Flow velocity fields surrounding the valve leaflets were calculated from PIV measurements and comparisons to the FSI simulation results were made. Both the spatial and temporal results of the simulations and experiments were in agreement. The FSI prediction of the transition point from equilibrium phase to valve-closing phase had a 7% delay compared to the PD measurements, while the PIV measurements matched the PD exactly. FSI predictions of reversed flow were within 10% compared to PD measurements. Stagnation or stasis regions were observed in both simulations and experiments. The pressure differential across the valve and associated forces on the leaflets from simulations showed the valve mechanism to be pressure driven. The flow velocity simulations were highly consistent with the experimental results. The FSI simulation and force analysis showed that the valve closure mechanism is pressure driven under the test conditions. FSI simulation and PIV measurements demonstrated that the flow behind the leaflet was mostly stagnant and a potential source for thrombosis. The validated FSI simulations should enable future valve design optimizations that are needed for

  12. Novel Active Combustion Control Valve

    Science.gov (United States)

    Caspermeyer, Matt

    2014-01-01

    This project presents an innovative solution for active combustion control. Relative to the state of the art, this concept provides frequency modulation (greater than 1,000 Hz) in combination with high-amplitude modulation (in excess of 30 percent flow) and can be adapted to a large range of fuel injector sizes. Existing valves often have low flow modulation strength. To achieve higher flow modulation requires excessively large valves or too much electrical power to be practical. This active combustion control valve (ACCV) has high-frequency and -amplitude modulation, consumes low electrical power, is closely coupled with the fuel injector for modulation strength, and is practical in size and weight. By mitigating combustion instabilities at higher frequencies than have been previously achieved (approximately 1,000 Hz), this new technology enables gas turbines to run at operating points that produce lower emissions and higher performance.

  13. Heavy gas valves

    Energy Technology Data Exchange (ETDEWEB)

    Steier, L [Vereinigte Armaturen Gesellschaft m.b.H., Mannheim (Germany, F.R.)

    1979-01-01

    Heavy gas valves must comply with special requirements. Apart from absolute safety in operation there are stringent requirements for material, sealing and ease of operation even in the most difficult conditions. Ball valves and single plate pipe gate valves lateral sealing rings have a dual, double sided sealing effect according to the GROVE sealing system. Single plate gate valves with lateral protective plates are suitable preferably for highly contaminated media. Soft sealing gate valves made of cast iron are used for low pressure applications.

  14. Relief valve testing study

    International Nuclear Information System (INIS)

    BROMM, R.D.

    2001-01-01

    Reclosing pressure-actuated valves, commonly called relief valves, are designed to relieve system pressure once it reaches the set point of the valve. They generally operate either proportional to the differential between their set pressure and the system pressure (gradual lift) or by rapidly opening fully when the set pressure is reached (pop action). A pop action valve allows the maximum fluid flow through the valve when the set pressure is reached. A gradual lift valve allows fluid flow in proportion to how much the system pressure has exceeded the set pressure of the valve (in the case of pressure relief) or has decreased below the set pressure (vacuum relief). These valves are used to protect systems from over and under pressurization. They are used on boilers, pressure vessels, piping systems and vacuum systems to prevent catastrophic failures of these systems, which can happen if they are under or over pressurized beyond the material tolerances. The construction of these valves ranges from extreme precision of less than a psi tolerance and a very short lifetime to extremely robust construction such as those used on historic railroad steam engines that are designed operate many times a day without changing their set pressure when the engines are operating. Relief valves can be designed to be immune to the effects of back pressure or to be vulnerable to it. Which type of valve to use depends upon the design requirements of the system

  15. Animal experimental studies on the influence of flow restriction on thrombogenicity of the Palmaz stent via 111indium marked thrombocytes

    International Nuclear Information System (INIS)

    Noeldge, G.; Siegerstetter, V.; Richter, G.M.; Garcia, O.; Palmaz, J.C.

    1990-01-01

    The purpose of this experimental study in dogs is to evaluate the influence of restricted flow on the thrombogenicity of balloon-expandable intravascular stents (Palmaz type). We implanted 24 Palmaz stents in the femoral artery of mongrel dogs. 12 dogs experienced a 75% flow restriction by means of an artificial stenosis distal to the start in the outflow tract. We used 111 In labelled thrombocytes as a marker for thrombus detection within the stent. The results presented demonstrate that there is no considerable risk of thrombus formation if the stent is implanted into vessels with good outflow tracts and therefore high blood flow velocities within the stent. Under conditions of highly impaired flow, a 68% rate of stent occlusion occur if no anticoagulation therapy is performed. After i.v. administration of 100 IU/kg body weight of heparin the occlusion rade drops to zero. It can be concluded that even under low-flow conditions as a result of a restricted outflow tract situation the Palmaz stent can be implanted without any risk of stent occlusion as long a sufficient anticoagulation protocol is maintained. (orig.) [de

  16. Recent experience with testing of parallel disc gate valves under accident flow conditions

    International Nuclear Information System (INIS)

    LaPointe, P.A.; Clayton, J.K.

    1992-01-01

    This paper presents the nuclear valve industry's latest and most extensive valve qualification test program experience. The test program includes a variety of 25 different gate and globe valves. All the test valves are power operated using either air, electric, or gas/hydraulic operators. The valves are categorized in size and pressure class so as to form a group of appropriate parent valve assemblies. Parent valve assembly qualification is used as the basis for qualification of candidate valve assemblies. The parent and candidate valve assemblies are representative of a nuclear plant's safety-related valve applications. The test program was performed in accordance with ANSI B16.41-1983 'Functional Qualification Requirements for Power Operated Active Valve Assemblies for Nuclear Power Plants.' The focus of this paper is on functional valve qualification test experience and specifically flow interruption testing to Annex G of the aforementioned test standard. Results of the flow test are summarized, including the coefficient of friction for each of the gate type valves reported. Information on valve size, pressure class, and actuator are given for all valves in the program. Although all valves performed extremely well, only selected test data are presented. The effects of the speed of operation and the effects of different fluid flow rates as they relate to the coefficient of friction between the valve disc and seat are discussed. The variation in the coefficient of friction based on other variables in the thrust equation, namely, differential pressure area is cited

  17. Experience with control valve cavitation problems and their solutions

    International Nuclear Information System (INIS)

    Ozol, J.

    1988-01-01

    Pressure reduction in control valves can induce cavitation, which has three effects on the control valve. Firstly, it modifies or changes the hydraulic performance of the control valve. Since control valves are designed for noncavitating conditions, the result is usually reduced stability of the control valve or, in extreme cavitating conditions known as supercavitation, the valve may limit the flow rate and thus be undersized. Secondly, cavitation can cause material damage to valve parts, trim, or valve body, or erodes downstream piping; consequently, the valve or piping leaks. Thirdly, cavitation causes noise and vibration, which may cause major damage or destruction to equipment such as valve positioners, actuators, pipe supports and sometimes to other downstream valves. The purpose of this paper is twofold: (1) It describes the I.S.A. valve sizing equations and how they relate to cavitation. (2) It describes experiences with these three problems, and discusses corrective actions and practical approaches to their solution. This paper discusses thirteen cavitation experiences

  18. An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves.

    Science.gov (United States)

    Ellis, J T; Healy, T M; Fontaine, A A; Weston, M W; Jarret, C A; Saxena, R; Yoganathan, A P

    1996-11-01

    indicate a propensity for blood element damage during the reverse flow phase of the cardiac cycle, but the extent of flow disturbance was twice as high with the Medtronic Parallel than with the St. Jude Medical valve. This research should yield a better understanding of the significance of retrograde flow to the functionality and potential thrombogenicity of bileaflet mechanical heart valves and aid in the development of new designs.

  19. Guide to prosthetic cardiac valves

    International Nuclear Information System (INIS)

    Morse, D.; Steiner, R.M.; Fernandez, J.

    1985-01-01

    This book contains 10 chapters. Some of the chapter titles are: The development of artificial heart valves: Introduction and historical perspective; The radiology of prosthetic heart valves; The evaluation of patients for prosthetic valve implantation; Pathology of cardiac valve replacement; and Bioengineering of mechanical and biological heart valve substitutes

  20. Heart valve replacement with the Sorin tilting-disc prosthesis. A 10-year experience.

    Science.gov (United States)

    Milano, A; Bortolotti, U; Mazzucco, A; Mossuto, E; Testolin, L; Thiene, G; Gallucci, V

    1992-02-01

    From 1978 to 1988, 697 patients with a mean age of 48 +/- 11 years (range 5 to 75 years) received a Sorin tilting-disc prosthesis; 358 had had aortic valve replacement, 247 mitral valve replacement, and 92 mitral and aortic valve replacement. Operative mortality rates were 7.8%, 11.3%, and 10.8%, respectively, in the three groups. Cumulative duration of follow-up is 1650 patient-years for aortic valve replacement (maximum follow-up 11.4 years), 963 patient-years for mitral valve replacement (maximum follow-up 9.9 years) and 328 patient-years for mitral and aortic valve replacement (maximum follow-up 9.4 years). Actuarial survival at 9 years is 72% +/- 4% after mitral valve replacement, 70% +/- 3% after aortic valve replacement, and 50% +/- 12% after mitral and aortic valve replacement, and actuarial freedom from valve-related deaths is 97% +/- 2% after mitral valve replacement, 92% +/- 2% after aortic valve replacement, and 62% +/- 15% after mitral and aortic valve replacement. Thromboembolic events occurred in 21 patients with aortic valve replacement (1.3% +/- 0.2%/pt-yr), in 12 with mitral valve replacement (1.2% +/- 0.3% pt-yr), and in seven with mitral and aortic valve replacement (2.1% +/- 0.8%), with one case of prosthetic thrombosis in each group; actuarial freedom from thromboembolism at 9 years is 92% +/- 3% after mitral valve replacement, 91% +/- 3% after aortic valve replacement, and 74% +/- 16% after mitral and aortic valve replacement. Anticoagulant-related hemorrhage was observed in 15 patients after aortic valve replacement (0.9% +/- 0.2%/pt-yr), in 9 after mitral valve replacement (0.9% +/- 0.3%/pt-yr), and in 6 with mitral and aortic valve replacement (0.9% +/- 0.5%/pt-yr); actuarial freedom from this complication at 9 years is 94% +/- 2% after aortic valve replacement, 91% +/- 4% after mitral valve replacement, and 68% +/- 16% after mitral and aortic valve replacement. Actuarial freedom from reoperation at 9 years is 97% +/- 2% after mitral and

  1. Intelligent Flow Control Valve

    Science.gov (United States)

    Kelley, Anthony R (Inventor)

    2015-01-01

    The present invention is an intelligent flow control valve which may be inserted into the flow coming out of a pipe and activated to provide a method to stop, measure, and meter flow coming from the open or possibly broken pipe. The intelligent flow control valve may be used to stop the flow while repairs are made. Once repairs have been made, the valve may be removed or used as a control valve to meter the amount of flow from inside the pipe. With the addition of instrumentation, the valve may also be used as a variable area flow meter and flow controller programmed based upon flowing conditions. With robotic additions, the valve may be configured to crawl into a desired pipe location, anchor itself, and activate flow control or metering remotely.

  2. Cryogenic Cam Butterfly Valve

    Science.gov (United States)

    McCormack, Kenneth J. (Inventor)

    2016-01-01

    A cryogenic cam butterfly valve has a body that includes an axially extending fluid conduit formed there through. A disc lug is connected to a back side of a valve disc and has a circular bore that receives and is larger than a cam of a cam shaft. The valve disc is rotatable for a quarter turn within the body about a lug axis that is offset from the shaft axis. Actuating the cam shaft in the closing rotational direction first causes the camming side of the cam of the cam shaft to rotate the disc lug and the valve disc a quarter turn from the open position to the closed position. Further actuating causes the camming side of the cam shaft to translate the valve disc into sealed contact with the valve seat. Opening rotational direction of the cam shaft reverses these motions.

  3. Low noise control valve

    International Nuclear Information System (INIS)

    Christie, R.S.

    1975-01-01

    Noise is one of the problems associated with the use of any type of control valve in systems involving the flow of fluids. The advent of OSHA standards has prompted control valve manufacturers to design valves with special trim to lower the sound pressure level to meet these standards. However, these levels are in some cases too high, particularly when a valve must be located in or near an area where people are working at tasks requiring a high degree of concentration. Such locations are found around and near research devices and in laboratory-office areas. This paper describes a type of fluid control device presently being used at PPL as a bypass control valve in deionized water systems and designed to reduce sound pressure levels considerably below OSHA standards. Details of the design and construction of this constant pressure drop variable flow control valve are contained in the text and are shown in photographs and drawings. Test data taken are included

  4. Valve packing manual. A maintenance application guide

    International Nuclear Information System (INIS)

    Aikin, J.A.; McCutcheon, R.G.; Cumming, D.

    1997-01-01

    Since 1970, AECL Chalk River Mechanical Equipment Development (MED) branch has invested over 175 person years in testing related to improving valve packing performance. Successful developments, including, 'live-loading', reduced packing heights, and performance-based packing qualification testing have been implemented. Since 1986, MED and the Integrated Valve Actuator Program Task Force - Valve Packing Steering Committee (IVAP-VPSC) have been involved in the development of combination die-formed graphite packing for use in CANDU plants. Many reports, articles, and specifications have been issued. Due to the large amount of test data and reports, a more user-friendly document has been prepared for everyday use. The Valve Packing Manual is based on many years of MED research and testing, as well as operating experience from CANDU nuclear generating stations (NGS). Since 1986, packing research and testing has been funded by the CANDU Owners Group (COG), the Electric Power Research Institute (EPRI), and participating valve packing manufacturers. The Valve Packing Manual (VPM) provides topical summaries of all work related to valve packing done since 1985. It includes advances in configuration design, stem packing friction, materials specifications, and installation procedures. This paper provides an overview on the application of the VPM with a focus on qualification testing, packing configuration, and stem packing friction. (author)

  5. Magnon Valve Effect between Two Magnetic Insulators

    Science.gov (United States)

    Wu, H.; Huang, L.; Fang, C.; Yang, B. S.; Wan, C. H.; Yu, G. Q.; Feng, J. F.; Wei, H. X.; Han, X. F.

    2018-03-01

    The key physics of the spin valve involves spin-polarized conduction electrons propagating between two magnetic layers such that the device conductance is controlled by the relative magnetization orientation of two magnetic layers. Here, we report the effect of a magnon valve which is made of two ferromagnetic insulators (YIG) separated by a nonmagnetic spacer layer (Au). When a thermal gradient is applied perpendicular to the layers, the inverse spin Hall voltage output detected by a Pt bar placed on top of the magnon valve depends on the relative orientation of the magnetization of two YIG layers, indicating the magnon current induced by the spin Seebeck effect at one layer affects the magnon current in the other layer separated by Au. We interpret the magnon valve effect by the angular momentum conversion and propagation between magnons in two YIG layers and conduction electrons in the Au layer. The temperature dependence of the magnon valve ratio shows approximately a power law, supporting the above magnon-electron spin conversion mechanism. This work opens a new class of valve structures beyond the conventional spin valves.

  6. Aortic or Mitral Valve Replacement With the Biocor and Biocor Supra

    Science.gov (United States)

    2017-04-26

    Aortic Valve Insufficiency; Aortic Valve Regurgitation; Aortic Valve Stenosis; Aortic Valve Incompetence; Mitral Valve Insufficiency; Mitral Valve Regurgitation; Mitral Valve Stenosis; Mitral Valve Incompetence

  7. Factors influencing mortality after bioprosthetic valve replacement; a midterm outcome.

    Science.gov (United States)

    Javadzadegan, Hassan; Javadzadegan, Amir; Mehdizadeh Baghbani, Jafar

    2013-01-01

    Although valve repair is applied routinely nowadays, particularly for mitral regurgitation (MR) or tricuspid regurgitation (TR), valve replacement using prosthetic valves is also common especially in adults. Unfortunately the valve with ideal hemodynamic performance and long-term durability without increasing the risk of bleeding due to long-term anticoagulant therapy has not been introduced. Therefore, patients and physicians must choose either bioprosthetic or mechanical valves. Currently, there is an increasing clinical trend of using bioprosthetic valves instead of mechanical valves even in young patients apparently because of their advantages. Seventy patients undergone valvular replacement using bioprosthetic valves were evaluated by ECG and Echocardiography to assess the rhythm and ejection fracture. Mean follow-up time was 33 months (min 9, max 92). Mortality rate was 25.9% (n=18) within 8 years of follow-up. Statistical analysis showed a significant relation between atrial fibrillation rhythm and mortality (P=0.02). Morbidities occurred in 30 patients (42.8%). Significant statistical relation was found between the morbidities and age over 65 years old (P=0.005). In follow-up period, 4 cases (5.7%) underwent re-operation due to global valve dysfunction. Our study shows that using biprosthetic valve could reduce the risk of morbidity occurrence in patient who needs valve replacement. However, if medical treatments fail, patients should be referred for surgery. This would reduce the risk of mortality because of lower incident of complications such as atrial fibrillation and morbidities due to younger patients' population.

  8. Magnetic Check Valve

    Science.gov (United States)

    Morris, Brian G.; Bozeman, Richard J., Jr.

    1994-01-01

    Poppet in proposed check valve restored to closed condition by magnetic attraction instead of spring force. Oscillations suppressed, with consequent reduction of wear. Stationary magnetic disk mounted just upstream of poppet, also containing magnet. Valve body nonmagnetic. Forward pressure or flow would push poppet away from stationary magnetic disk so fluid flows easily around poppet. Stop in valve body prevents poppet from being swept away. When flow stopped or started to reverse, magnetic attraction draws poppet back to disk. Poppet then engages floating O-ring, thereby closing valve and preventing reverse flow. Floating O-ring facilitates sealing at low loads.

  9. Butterfly valves for seawater

    International Nuclear Information System (INIS)

    Yamanaka, Katsuto

    1991-01-01

    Recently in thermal and nuclear power stations and chemical plants which have become large capacity, large quantity of cooling water is required, and mostly seawater is utilized. In these cooling water systems, considering thermal efficiency and economy, the pipings become complex, and various control functions are demanded. For the purpose, the installation of shut-off valves and control valves for pipings is necessary. The various types of valves have been employed, and in particular, butterfly valves have many merits in their function, size, structure, operation, maintenance, usable period, price and so on. The corrosion behavior of seawater is complicated due to the pollution of seawater, therefore, the environment of the valves used for seawater became severe. The structure and the features of the butterfly valves for seawater, the change of the structure of the butterfly valves for seawater and the checkup of the butterfly valves for seawater are reported. The corrosion of metallic materials is complicatedly different due to the locating condition of plants, the state of pipings and the condition of use. The corrosion countermeasures for butterfly valves must be examined from the synthetic viewpoints. (K.I.)

  10. Redo mitral valve surgery

    Directory of Open Access Journals (Sweden)

    Redoy Ranjan

    2018-03-01

    Full Text Available This study is based on the findings of a single surgeon’s practice of mitral valve replacement of 167 patients from April 2005 to June 2017 who developed symptomatic mitral restenosis after closed or open mitral commisurotomy. Both clinical and color doppler echocardiographic data of peri-operative and six months follow-up period were evaluated and compared to assess the early outcome of the redo mitral valve surgery. With male-female ratio of 1: 2.2 and after a duration of 6 to 22 years symptom free interval between the redo procedures, the selected patients with mitral valve restenosis undergone valve replacement with either mechanical valve in 62% cases and also tissue valve in 38% cases. Particular emphasis was given to separate the adhered pericardium from the heart completely to ameliorate base to apex and global contraction of the heart. Besides favorable post-operative clinical outcome, the echocardiographic findings were also encouraging as there was statistically significant increase in the mitral valve area and ejection fraction with significant decrease in the left atrial diameter, pressure gradient across the mitral valve and pulmonary artery systolic pressure. Therefore, in case of inevitable mitral restenosis after closed or open commisurotomy, mitral valve replacement is a promising treatment modality.

  11. Function analysis of steam isolation valves

    International Nuclear Information System (INIS)

    Persson, R.; Kilpi, K.; Noro, H.; Siikonen, T.; Sjoeberg, A.; Wallen, G.; Aakesson, H.

    1981-01-01

    Function analysis of system-medium-operated steam isolation valves has been the objective of the Swedish-Finnish IVLS project, the results of which are presented in this report. Theoretical models were to be verified against available experimental data, to some extent from the HDR blowdown experiments, which are part of a German reactor safety program. Finnish hydraulic measurements on a valve model (scale 1:2.15) have been performed to give complementary data. The analysis work has covered the thermal-hydraulic behaviour of steam isolation valves as well as phenomena related to structural mechanics. Work performed under contract with the Swedish Nuclear Power Inspectorate. (Author)

  12. Function analysis of steam isolation valves

    International Nuclear Information System (INIS)

    Persson, R.; Sjoeberg, A.; Aakesson, H.; Kilpi, K.; Noro, H.; Siikonen, T.; Wallen, G.

    1981-01-01

    Function analysis of system-medium-operated steam isolation valves has been the objective of the Swedish-Finnish IVLS project, the results of which are presented in this report. Theoretical models were to be verified against available experimental data, to some extent from the HDR blowdown experiments, which are part of a German reactor safety program. Finnish hydraulic measurements on a valve model (scale 1:2.15) have been performed to give complementary data. The analysis work has covered the thermal-hydraulic behaviour of steam isolation valves as well as phenomena related to structural mechanics. (Auth.)

  13. Echocardiographic Assessment of Heart Valve Prostheses

    Science.gov (United States)

    Sordelli, Chiara; Severino, Sergio; Ascione, Luigi; Coppolino, Pasquale; Caso, Pio

    2014-01-01

    Patients submitted to valve replacement with mechanical or biological prosthesis, may present symptoms related either to valvular malfunction or ventricular dysfunction from other causes. Because a clinical examination is not sufficient to evaluate a prosthetic valve, several diagnostic methods have been proposed to assess the functional status of a prosthetic valve. This review provides an overview of echocardiographic and Doppler techniques useful in evaluation of prosthetic heart valves. Compared to native valves, echocardiographic evaluation of prosthetic valves is certainly more complex, both for the examination and the interpretation. Echocardiography also allows discriminating between intra- and/or peri-prosthetic regurgitation, present in the majority of mechanical valves. Transthoracic echocardiography (TTE) requires different angles of the probe with unconventional views. Transesophageal echocardiography (TEE) is the method of choice in presence of technical difficulties. Three-dimensional (3D)-TEE seems to be superior to 2D-TEE, especially in the assessment of paravalvular leak regurgitation (PVL) that it provides improved localization and analysis of the PVL size and shape. PMID:28465917

  14. Frequency and Effect of Access-Related Vascular Injury and Subsequent Vascular Intervention After Transcatheter Aortic Valve Replacement

    DEFF Research Database (Denmark)

    Dencker, Ditte; Taudorf, Mikkel; Luk, N H Vincent

    2016-01-01

    patients underwent TAVR and 333 patients (94%) were treated by true percutaneous transfemoral approach. Of this latter group, 83 patients (25%) had an access-related vascular injury that was managed by the use of a covered self-expanding stent (n = 49), balloon angioplasty (n = 33), or by surgical...

  15. Danfos: Thermostatic Radiator Valves

    DEFF Research Database (Denmark)

    Gregersen, Niels; Oliver, James; Hjorth, Poul G.

    2000-01-01

    This problem deals with modelling the flow through a typical Danfoss thermostatic radiator valve.Danfoss is able to employ Computational Fluid Dynamics (CFD) in calculations of the capacity of valves, but an experienced engineer can often by rules of thumb "guess" the capacity, with a precision...

  16. Bentall Procedure Using Cryopreserved Valved Aortic Homografts

    Science.gov (United States)

    Christenson, Jan T.; Sierra, Jorge; Trindade, Pedro T.; Didier, Dominique; Kalangos, Afksendiyos

    2004-01-01

    The Bentall procedure is the standard operation for patients who have lesions of the ascending aorta associated with aortic valve disease. In many cases, however, mechanical prosthetic conduits are not suitable. There are few reports in the English-language medical literature concerning the mid- to long-term outcome of Bentall operations with cryopreserved homografts. Therefore, we reviewed our experience with this procedure and valved homografts. From January 1997 through December 2002, 21 patients underwent a Bentall operation with cryopreserved homografts at our institution. There were 14 males and 7 females; the mean age was 36 ± 21 years (range, 15–74 years). Eleven patients had undergone previous aortic valve surgery. All patients had aortic dilatation or aneurysms involving the ascending aorta. Indications for surgery included aortic valve stenosis or insufficiency, and aortic valve endocarditis (native valve or prosthetic). One patient had Takayasu's arteritis and 3 had Marfan syndrome. There was 1 hospital death (due to sepsis), but no other major postoperative complications. The mean hospital stay was 14 ± 7 days. Follow-up echocardiographic and computed tomographic scans were performed yearly. The mean follow-up was 34 months (6–72 months). Follow-up imaging revealed no calcifications or degenerative processes related to the homograft. Four patients had minimal valve regurgitation. Two patients died during follow-up. The 3-year actuarial survival rate was 85.7%. Our data suggest that the Bentall procedure with a valved homograft conduit is a safe procedure with excellent mid- to long-term results, comparable to results reported with aortic valve replacement with a homograft. PMID:15745290

  17. Mitral valve prolapse and hyperthyroidism: effect of patient selection.

    Science.gov (United States)

    Zullo, M A; Devereux, R B; Kramer-Fox, R; Lutas, E M; Brown, W T

    1985-11-01

    Patients with mitral valve prolapse and hyperthyroidism have common symptoms; the most outstanding symptom is palpitation. To determine whether or not common symptoms contributed to the reported association of these conditions, we evaluated 220 patients with symptomatic mitral valve prolapse and 216 first-degree relatives in 72 families; 65 relatives with mitral valve prolapse and 151 relatives without mitral valve prolapse, all greater than or equal to 16 years of age. Thirty subjects, aged 49 +/- 13 years (p less than 0.025 vs entire study group), had thyroid disease (23 subjects had definite thyroid disease, seven subjects had probable); 27 of 30 subjects with thyroid disease (90%) were female (p less than 0.005). The age- and sex-adjusted prevalence of hyperthyroidism was significantly higher in probands with mitral valve prolapse than in family members without mitral valve prolapse (3.5% vs 0%, p = 0.03), while an intermediate prevalence of hyperthyroidism (2.2%) was observed in family members with mitral valve prolapse. Thus, the prevalence of hyperthyroidism is increased among symptomatic patients with mitral valve prolapse as compared to family members without mitral valve prolapse, but the prevalence of thyroid conditions is similar among family members with or without this condition. These findings are explained by the effect of common symptoms on clinical detection of both mitral valve prolapse and hyperthyroidism.

  18. Bioprinting a cardiac valve.

    Science.gov (United States)

    Jana, Soumen; Lerman, Amir

    2015-12-01

    Heart valve tissue engineering could be a possible solution for the limitations of mechanical and biological prostheses, which are commonly used for heart valve replacement. In tissue engineering, cells are seeded into a 3-dimensional platform, termed the scaffold, to make the engineered tissue construct. However, mimicking the mechanical and spatial heterogeneity of a heart valve structure in a fabricated scaffold with uniform cell distribution is daunting when approached conventionally. Bioprinting is an emerging technique that can produce biological products containing matrix and cells, together or separately with morphological, structural and mechanical diversity. This advance increases the possibility of fabricating the structure of a heart valve in vitro and using it as a functional tissue construct for implantation. This review describes the use of bioprinting technology in heart valve tissue engineering. Copyright © 2015 Elsevier Inc. All rights reserved.

  19. Space Vehicle Valve System

    Science.gov (United States)

    Kelley, Anthony R. (Inventor); Lindner, Jeffrey L. (Inventor)

    2014-01-01

    The present invention is a space vehicle valve system which controls the internal pressure of a space vehicle and the flow rate of purged gases at a given internal pressure and aperture site. A plurality of quasi-unique variable dimension peaked valve structures cover the purge apertures on a space vehicle. Interchangeable sheet guards configured to cover valve apertures on the peaked valve structure contain a pressure-activated surface on the inner surface. Sheet guards move outwardly from the peaked valve structure when in structural contact with a purge gas stream flowing through the apertures on the space vehicle. Changing the properties of the sheet guards changes the response of the sheet guards at a given internal pressure, providing control of the flow rate at a given aperture site.

  20. Multiple-port valve

    International Nuclear Information System (INIS)

    Doody, T.J.

    1978-01-01

    A multiple-port valve assembly is designed to direct flow from a primary conduit into any one of a plurality of secondary conduits as well as to direct a reverse flow. The valve includes two mating hemispherical sockets that rotatably receive a spherical valve plug. The valve plug is attached to the primary conduit and includes diverging passageways from that conduit to a plurality of ports. Each of the ports is alignable with one or more of a plurality of secondary conduits fitting into one of the hemispherical sockets. The other hemispherical socket includes a slot for the primary conduit such that the conduit's motion along that slot with rotation of the spherical plug about various axes will position the valve-plug ports in respect to the secondary conduits

  1. Volume-constrained optimization of magnetorheological and electrorheological valves and dampers

    Science.gov (United States)

    Rosenfeld, Nicholas C.; Wereley, Norman M.

    2004-12-01

    This paper presents a case study of magnetorheological (MR) and electrorheological (ER) valve design within a constrained cylindrical volume. The primary purpose of this study is to establish general design guidelines for volume-constrained MR valves. Additionally, this study compares the performance of volume-constrained MR valves against similarly constrained ER valves. Starting from basic design guidelines for an MR valve, a method for constructing candidate volume-constrained valve geometries is presented. A magnetic FEM program is then used to evaluate the magnetic properties of the candidate valves. An optimized MR valve is chosen by evaluating non-dimensional parameters describing the candidate valves' damping performance. A derivation of the non-dimensional damping coefficient for valves with both active and passive volumes is presented to allow comparison of valves with differing proportions of active and passive volumes. The performance of the optimized MR valve is then compared to that of a geometrically similar ER valve using both analytical and numerical techniques. An analytical equation relating the damping performances of geometrically similar MR and ER valves in as a function of fluid yield stresses and relative active fluid volume, and numerical calculations are provided to calculate each valve's damping performance and to validate the analytical calculations.

  2. The prognosis of infective endocarditis treated with biological valves versus mechanical valves: A meta-analysis.

    Science.gov (United States)

    Tao, Ende; Wan, Li; Wang, WenJun; Luo, YunLong; Zeng, JinFu; Wu, Xia

    2017-01-01

    receiving a biological valve compared to patients receiving a mechanical valve. A large, multicenter retrospective study included in our meta-analysis suggested that any mortality risk of the biological valve group was significant higher than that of the mechanical valve group. However, the risk was no different after risk was adjusted. So, we thought the reason for this result may be related to the characteristics of the patient rather than valve dysfunction. It is still necessary to future randomized studies to verify this conclusion.

  3. 19 mm sized bileaflet valve prostheses' flow field investigated by bidimensional laser Doppler anemometry (part II: maximum turbulent shear stresses)

    Science.gov (United States)

    Barbaro, V; Grigioni, M; Daniele, C; D'Avenio, G; Boccanera, G

    1997-11-01

    The investigation of the flow field generated by cardiac valve prostheses is a necessary task to gain knowledge on the possible relationship between turbulence-derived stresses and the hemolytic and thrombogenic complications in patients after valve replacement. The study of turbulence flows downstream of cardiac prostheses, in literature, especially concerns large-sized prostheses with a variable flow regime from very low up to 6 L/min. The Food and Drug Administration draft guidance requires the study of the minimum prosthetic size at a high cardiac output to reach the maximum Reynolds number conditions. Within the framework of a national research project regarding the characterization of cardiovascular endoprostheses, an in-depth study of turbulence generated downstream of bileaflet cardiac valves is currently under way at the Laboratory of Biomedical Engineering of the Istituto Superiore di Sanita. Four models of 19 mm bileaflet valve prostheses were used: St Jude Medical HP, Edwards Tekna, Sorin Bicarbon, and CarboMedics. The prostheses were selected for the nominal Tissue Annulus Diameter as reported by manufacturers without any assessment of valve sizing method, and were mounted in aortic position. The aortic geometry was scaled for 19 mm prostheses using angiographic data. The turbulence-derived shear stresses were investigated very close to the valve (0.35 D0), using a bidimensional Laser Doppler anemometry system and applying the Principal Stress Analysis. Results concern typical turbulence quantities during a 50 ms window at peak flow in the systolic phase. Conclusions are drawn regarding the turbulence associated to valve design features, as well as the possible damage to blood constituents.

  4. Characterization of the thrombogenic potential of surface oxides on stainless steel for implant purposes

    International Nuclear Information System (INIS)

    Shih, C.-C.; Shih, C.-M.; Su, Y.-Y.; Chang, M.-S.; Lin, S.-J.

    2003-01-01

    Marketed stents are manufactured from various metals and passivated with different degrees of surface oxidation. The functional surface oxides on the degree of antithrombotic potential were explored through a canine femoral extracorporeal circuit model. Related properties of these oxide films were studied by open-circuit potential, current density detected at open-circuit potential, the electrochemical impedance spectroscopy, transmission electron microscopy, Auger spectroscopy (AES), X-ray photoelectron spectroscopy (XPS), and scanning electron microscopy. Experimental evidences showed that blood clot weight after a 30-min follow-up was significantly lower for the stainless steel wire passivated with amorphous oxide (AO) compared to the wire passivated with polycrystalline oxide (PO) or commercial as-received wire coils (AS). Surface characterizations showed that a stable negative current density at open-circuit potential and a significant lower potential were found for the wire surface passivated with AO than for the surface passivated with PO. Time constant of AO is about 25 times larger than that of polycrystalline oxide. Significant difference in oxide grain sizes was found between PO and AO. Surface chemistries revealed by the AES and XPS spectra indicated the presence of a Cr- and oxygen-rich surface oxide for AO, and a Fe-rich and oxygen-lean surface oxide for PO. These remarkable characteristics of AO surface film might have a potential to provide for excellent antithrombotic characteristics for the 316L stainless steel stents

  5. Human plasma-derived immunoglobulin G fractionated by an aqueous two-phase system, caprylic acid precipitation, and membrane chromatography has a high purity level and is free of detectable in vitro thrombogenic activity.

    Science.gov (United States)

    Vargas, M; Segura, Á; Wu, Y-W; Herrera, M; Chou, M-L; Villalta, M; León, G; Burnouf, T

    2015-02-01

    Instituto Clodomiro Picado has developed an immunoglobulin G (IgG) plasma fractionation process combining a polyethylene glycol/phosphate aqueous two-phase system (ATPS), caprylic acid precipitation and anion-exchange membrane chromatography. We evaluated the purity and in vitro thrombogenicity of such IgG, in line with current international requirements. Contributions of the different production steps to reduce thrombogenicity were assessed at 0·2 l-scale, and then the methodology was scaled-up to a 10 l-scale and final products (n = 3) were analysed. Purity, immunoglobulin composition, and subclass distribution were determined by electrophoretic and immunochemical methods. The in vitro thrombogenic potential was determined by a thrombin generation assay (TGA) using a Technothrombin fluorogenic substrate. Prekallikrein activator (PKA), plasmin, factor Xa, thrombin and thrombin-like activities were assessed using S-2302, S-2251, S-2222, S-2238 and S-2288 chromogenic substrates, respectively, and FXI by an ELISA. The thrombogenicity markers were reduced mostly during the ATPS step and were found to segregate mostly into the discarded liquid upper phase. The caprylic acid precipitation eliminated the residual procoagulant activity. The IgG preparations made from the 10 l-batches contained 100% gamma proteins, low residual IgA and undetectable IgM. The IgG subclass distribution was not substantially affected by the process. TGA and amidolytic activities revealed an undetectable in vitro thrombogenic risk and the absence of proteolytic enzymes in the final product. Fractionating human plasma by an ATPS combined with caprylic acid and membrane chromatography resulted in an IgG preparation of high purity and free of a detectable in vitro thrombogenic risk. © 2014 International Society of Blood Transfusion.

  6. Gate valve performance prediction

    International Nuclear Information System (INIS)

    Harrison, D.H.; Damerell, P.S.; Wang, J.K.; Kalsi, M.S.; Wolfe, K.J.

    1994-01-01

    The Electric Power Research Institute is carrying out a program to improve the performance prediction methods for motor-operated valves. As part of this program, an analytical method to predict the stem thrust required to stroke a gate valve has been developed and has been assessed against data from gate valve tests. The method accounts for the loads applied to the disc by fluid flow and for the detailed mechanical interaction of the stem, disc, guides, and seats. To support development of the method, two separate-effects test programs were carried out. One test program determined friction coefficients for contacts between gate valve parts by using material specimens in controlled environments. The other test program investigated the interaction of the stem, disc, guides, and seat using a special fixture with full-sized gate valve parts. The method has been assessed against flow-loop and in-plant test data. These tests include valve sizes from 3 to 18 in. and cover a considerable range of flow, temperature, and differential pressure. Stem thrust predictions for the method bound measured results. In some cases, the bounding predictions are substantially higher than the stem loads required for valve operation, as a result of the bounding nature of the friction coefficients in the method

  7. Modeling valve leakage

    International Nuclear Information System (INIS)

    Bell, S.R.; Rohrscheib, R.

    1994-01-01

    The American Society of Mechanical Engineers (ASME) Code requires individual valve leakage testing for Category A valves. Although the U.S. Nuclear Regulatory Commission (USNRC) has recognized that it is more appropriate to test containment isolation valves in groups, as allowed by 10 CFR 50, Appendix J, a utility seeking relief from these Code requirements must provide technical justification for the relief and establish a conservative alternate acceptance criteria. In order to provide technical justification for group testing of containment isolation valves, Illinois Power developed a calculation (model) for determining the size of a leakage pathway in a valve disc or seat for a given leakage rate. The model was verified experimentally by machining leakage pathways of known size and then measuring the leakage and comparing this value to the calculated value. For the range of values typical of leakage rate testing, the correlation between the experimental values and calculated values was quote good. Based upon these results, Illinois Power established a conservative acceptance criteria for all valves in the inservice testing (IST) program and was granted relief by the USNRC from the individual leakage testing requirements of the ASME Code. This paper presents the results of Illinois Power's work in the area of valve leakage rate testing

  8. Face-Sealing Butterfly Valve

    Science.gov (United States)

    Tervo, John N.

    1992-01-01

    Valve plate made to translate as well as rotate. Valve opened and closed by turning shaft and lever. Interactions among lever, spring, valve plate, and face seal cause plate to undergo combination of translation and rotation so valve plate clears seal during parts of opening and closing motions.

  9. Relation of Aortic Valve and Coronary Artery Calcium in Patients With Chronic Kidney Disease to the Stage and Etiology of the Renal Disease

    NARCIS (Netherlands)

    Piers, Lieuwe H.; Touw, Hugo R. W.; Gansevoort, Ron; Franssen, Casper F. M.; Oudkerk, Matthijs; Zijlstra, Felix; Tio, Rene A.

    2009-01-01

    Patients with chronic renal failure have increased cardiac calcium loads. Previous studies have investigated the prevalence and quantitative extent of aortic valve calcium (AVC) and coronary artery calcium (CAC) in patients with various stages of chronic kidney disease (CKD). However, the impact of

  10. GIANT PROSTHETIC VALVE THROMBUS

    Directory of Open Access Journals (Sweden)

    Prashanth Kumar

    2015-04-01

    Full Text Available Mechanical prosthetic valves are predisposed to bleeding, thrombosis & thromboembolic complications. Overall incidence of thromboembolic complications is 1% per year who are on oral anticoagulants, whereas bleeding complications incidence is 0.5% to 6.6% per year. 1, 2 Minimization of Scylla of thromboembolic & Charybdis of bleeding complication needs a balancing act of optimal antithrombotic therapy. We are reporting a case of middle aged male patient with prosthetic mitral valve presenting in heart failure. Patient had discontinued anticoagulants, as he had subdural hematoma in the past. He presented to our institute with a giant prosthetic valve thrombus.

  11. Thrombogenicity and central pulse pressure to enhance prediction of ischemic event occurrence in patients with established coronary artery disease: The MAGMA-ischemia score.

    Science.gov (United States)

    Bliden, Kevin P; Chaudhary, Rahul; Navarese, Eliano P; Sharma, Tushar; Kaza, Himabindu; Tantry, Udaya S; Gurbel, Paul A

    2018-01-01

    Conventional cardiovascular risk estimators based on clinical demographics have limited prediction of coronary events. Markers for thrombogenicity and vascular function have not been explored in risk estimation of high-risk patients with coronary artery disease. We aimed to develop a clinical and biomarker score to predict 3-year adverse cardiovascular events. Four hundred eleven patients, with ejection fraction ≥40% undergoing coronary angiography, and found to have a luminal diameter stenosis ≥50%, were included in the analysis. Thrombelastography indices and central pulse pressure (CPP) were determined at the time of catheterization. We identified predictors of death, myocardial infarction (MI) or stroke and developed a numerical ischemia risk score. The primary endpoint of cardiovascular death, MI or stroke occurred in 22 patients (5.4%). The factors associated with events were age, prior PCI or CABG, diabetes, CPP, and thrombin-induced platelet-fibrin clot strength, and were included in the MAGMA-ischemia score. The MAGMA-ischemia score showed a c-statistic of 0.85 (95% Confidence Interval [CI] 0.80-0.87; pMAGMA-ischemia score greater than 5 had highest risk to develop clinical events, hazard ratio for the primary endpoint: 13.9 (95% CI 5.8-33.1, pMAGMA-ischemia score yielded a higher discrimination. Inclusion of CPP and assessment of thrombogenicity in a novel score for patients with documented CAD enhanced the prediction of events. Copyright © 2017 Elsevier B.V. All rights reserved.

  12. Virgin olive oil, palm olein and coconut oil diets do not raise cell adhesion molecules and thrombogenicity indices in healthy Malaysian adults.

    Science.gov (United States)

    Voon, P T; Ng, T K W; Lee, V K M; Nesaretnam, K

    2015-06-01

    Effects of high-protein diets that are rich in saturated fats on cell adhesion molecules, thrombogenicity and other nonlipid markers of atherosclerosis in humans have not been firmly established. We aim to investigate the effects of high-protein Malaysian diets prepared separately with virgin olive oil (OO), palm olein (PO) and coconut oil (CO) on cell adhesion molecules, lipid inflammatory mediators and thromobogenicity indices in healthy adults. A randomized cross-over intervention with three dietary sequences, using virgin OO, PO and CO as test fats, was carried out for 5 weeks on each group consisting of 45 men and women. These test fats were incorporated separately at two-thirds of 30% fat calories into high-protein Malaysian diets. For fasting and nonfasting blood samples, no significant differences were observed on the effects of the three test-fat diets on thrombaxane B2 (TXB2), TXB2/PGF1α ratios and soluble intracellular and vascular cell adhesion molecules. The OO diet induced significantly lower (Pvirgin OO do not alter the thrombogenicity indices-cellular adhesion molecules, thromboxane B2 (TXB2) and TXB2/prostacyclin (PGF1α) ratios. However, the OO diet lowered plasma proinflammatory LTB4, whereas the PO diet raised the antiaggregatory plasma PGF1α in healthy Malaysian adults. This trial was registered at clinicaltrials.gov as NCT 00941837.

  13. Valve monitoring ITI-MOVATS

    International Nuclear Information System (INIS)

    Moureau, S.

    1993-01-01

    ITI-MOVATS provides a wide range of test devices to monitor the performance of valves: motor operated gate or globe valve, butterfly valve, air operated valve, and check valve. The ITI-MOVATS testing equipment is used in the following three areas: actuator setup/baseline testing, periodic/post-maintenance testing, and differential pressure testing. The parameters typically measured with the MOVATS diagnostic system as well as the devices used to measure them are described. (Z.S.)

  14. Bioprosthetic Valve Fracture to Facilitate Transcatheter Valve-in-Valve Implantation.

    Science.gov (United States)

    Allen, Keith B; Chhatriwalla, Adnan K; Cohen, David J; Saxon, John T; Aggarwal, Sanjeev; Hart, Anthony; Baron, Suzanne; Davis, J Russell; Pak, Alex F; Dvir, Danny; Borkon, A Michael

    2017-11-01

    Valve-in-valve transcatheter aortic valve replacement is less effective in small surgical bioprostheses. We evaluated the feasibility of bioprosthetic valve fracture with a high-pressure balloon to facilitate valve-in-valve transcatheter aortic valve replacement. In vitro bench testing on aortic tissue valves was performed on 19-mm and 21-mm Mitroflow (Sorin, Milan, Italy), Magna and Magna Ease (Edwards Lifesciences, Irvine, CA), Trifecta and Biocor Epic (St. Jude Medical, Minneapolis, MN), and Hancock II and Mosaic (Medtronic, Minneapolis, MN). High-pressure balloons Tru Dilation, Atlas Gold, and Dorado (C.R. Bard, Murray Hill, NJ) were used to determine which valves could be fractured and at what pressure fracture occurred. Mitroflow, Magna, Magna Ease, Mosaic, and Biocor Epic surgical valves were successfully fractured using high-pressures balloon 1 mm larger than the labeled valve size whereas Trifecta and Hancock II surgical valves could not be fractured. Only the internal valve frame was fractured, and the sewing cuff was never disrupted. Manufacturer's rated burst pressures for balloons were exceeded, with fracture pressures ranging from 8 to 24 atmospheres depending on the surgical valve. Testing further demonstrated that fracture facilitated the expansion of previously constrained, underexpanded transcatheter valves (both balloon and self-expanding) to the manufacturer's recommended size. Bench testing demonstrates that the frame of most, but not all, bioprosthetic surgical aortic valves can be fractured using high-pressure balloons. The safety of bioprosthetic valve fracture to optimize valve-in-valve transcatheter aortic valve replacement in small surgical valves requires further clinical investigation. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  15. NRC perspective and experience on valve testing

    International Nuclear Information System (INIS)

    Eapen, P.K.

    1990-01-01

    Testing of safety related valves is one of the major activities at commercial nuclear power plants. In addition to Technical Specification, valve testing is required in 10 CFR 50.55a and 10 CFR 50 Appendix J. NRC inspectors (both resident and specialists) spend a considerable amount of time in following the valve test activities as part of their routine business. In the past, depending on a licensee's organizational structure, a valve could be tested more than three times to verify conformance with Technical Specifications, 10 CFR 50.55a, and 10 CFR 50 Appendix J. The regulatory reviewers were isolated from each other. Licensee test personnel were also not communicating among themselves. As a result, NRC inspectors found that certain valves in the IST program were inadequately tested. The typical licensee response was to say that this valve is exempted from testing under Appendix J. Others would say that the technical specification does not require fast closure of a valve in question. In addition to the above, the inspectors had to deal with exemption requests that were not dispositioned by the NRC. In the seventies there was a gentlemen's agreement to allow the licensee to do the testing in accordance with the exception, without waiting for the NRC approval. Needless to say when the new NRC inspection procedure was issued in March 1989 for implementation, the Regional inspectors had extremely difficult time to cope with the gray areas of valve testing. In August 1987, NRC Region I was reorganized and the special test program section was established to perform inspections in the IST area. This section was chartered to optimize resources and develop a meaningful inspection plan. The perspectives and insights used in the development of a detailed inspection plan is discussed below

  16. Improved corrosion resistance of spin-valve film

    International Nuclear Information System (INIS)

    Tetsukawa, H.; Hommura, H.; Okabe, A.; Soda, Y.

    2007-01-01

    We investigated the corrosion behavior and magnetoresistance of spin-valve film in order to improve the corrosion resistance of the spin-valve head for a tape recording system. The conventional spin-valve head (sub./Ta/NiFe/CoFe/Cu/CoFe/PtMn/Ta) with no diamond-like carbon (DLC) protective layer showed poor corrosion resistance. This is because the CoFe for ferromagnetic layer and Cu for spacer in the spin-valve film exhibited poor corrosion resistance. The corrosion resistance of the CoFe film and Cu film improved with the addition of Ni and Au, respectively. The spin-valve film (sub./Ta/NiFe/CoNiFe/CuAu/CoNiFe/PtMn/Ta) showed higher pitting potential than the conventional spin-valve film by +0.45 V. This presents a significant improvement over the conventional spin-valve film. We also investigated the effect of the composition of ferromagnetic layer and spacer on the magnetoresistance of the spin-valve film. The magnetoresistance of the spin-valve film by substitution of CoNiFe for CoFe in ferromagnetic layer decreased slightly. The magnetoresistance of the spin-valve film decreased as the addition of Au of the spacer increased. The diffusion at CoNiFe/CuAu interface has not been observed in annealing process. The quantitative relation between corrosion resistance and magnetoresistance of spin-valve film, and its ferromagnetic layer and spacer's compositions have been clarified. The output voltage at 50 Oe of the corrosion-resistant spin-valve head with CoNiFe ferromagnetic layer and CuAu spacer was about 50% of that of the conventional spin-valve head

  17. Improved corrosion resistance of spin-valve film

    Energy Technology Data Exchange (ETDEWEB)

    Tetsukawa, H. [Sony Corporation, 6-7-35 Kitashinagawa, Shinagawa-ku, Tokyo 141-0001 (Japan)]. E-mail: tetsukaw@arc.sony.co.jp; Hommura, H. [Sony Corporation, 6-7-35 Kitashinagawa, Shinagawa-ku, Tokyo 141-0001 (Japan); Okabe, A. [Sony Corporation, 6-7-35 Kitashinagawa, Shinagawa-ku, Tokyo 141-0001 (Japan); Soda, Y. [Sony Corporation, 6-7-35 Kitashinagawa, Shinagawa-ku, Tokyo 141-0001 (Japan)

    2007-06-15

    We investigated the corrosion behavior and magnetoresistance of spin-valve film in order to improve the corrosion resistance of the spin-valve head for a tape recording system. The conventional spin-valve head (sub./Ta/NiFe/CoFe/Cu/CoFe/PtMn/Ta) with no diamond-like carbon (DLC) protective layer showed poor corrosion resistance. This is because the CoFe for ferromagnetic layer and Cu for spacer in the spin-valve film exhibited poor corrosion resistance. The corrosion resistance of the CoFe film and Cu film improved with the addition of Ni and Au, respectively. The spin-valve film (sub./Ta/NiFe/CoNiFe/CuAu/CoNiFe/PtMn/Ta) showed higher pitting potential than the conventional spin-valve film by +0.45 V. This presents a significant improvement over the conventional spin-valve film. We also investigated the effect of the composition of ferromagnetic layer and spacer on the magnetoresistance of the spin-valve film. The magnetoresistance of the spin-valve film by substitution of CoNiFe for CoFe in ferromagnetic layer decreased slightly. The magnetoresistance of the spin-valve film decreased as the addition of Au of the spacer increased. The diffusion at CoNiFe/CuAu interface has not been observed in annealing process. The quantitative relation between corrosion resistance and magnetoresistance of spin-valve film, and its ferromagnetic layer and spacer's compositions have been clarified. The output voltage at 50 Oe of the corrosion-resistant spin-valve head with CoNiFe ferromagnetic layer and CuAu spacer was about 50% of that of the conventional spin-valve head.

  18. Pulmonary valve stenosis

    Science.gov (United States)

    ... surgery - discharge Images Heart valves References Carabello BA. Valvular heart disease. In: Goldman L, Schafer AI, eds. Goldman's Cecil ... Saunders; 2016:chap 69. Otto CM, Bownow RO. Valvular heart disease. In: Mann DL, Zipes DP, Libby P, Bonow ...

  19. Mitral valve regurgitation

    Science.gov (United States)

    ... and dentist if you have a history of heart valve disease or congenital heart disease before treatment. Some people ... the middle Heart, front view References Carabello BA. Valvular heart disease. In: Goldman L, Schafer AI, eds. Goldman-Cecil ...

  20. Aortic Valve Disease

    Science.gov (United States)

    ... team will discuss with you the advantages and disadvantages of both valve types. Regardless of which type ... Diagnosis and Treatment Options Recovery Questions for Your Doctor Will my condition ever get better without treatment? ...

  1. Dry product valve

    International Nuclear Information System (INIS)

    Greaves, James D.

    1984-01-01

    This invention provides a system for delivering particulate radioactive or other toxic wastes to a container in which they can be solidified. The system includes a set of valves that prevent the escape of dusty materials to the atmosphere

  2. Ball check valve

    International Nuclear Information System (INIS)

    Bevilacqua, F.

    1978-01-01

    A pressurized nuclear reactor having an instrument assembly sheathed in a metallic tube which is extended vertically upward into the reactor core by traversing a metallic guide tube which is welded to the wall of the vessel is described. Sensors in each instrument assembly are connected to instruments outside the vessel to manifest the conditions within the core. Each instrument assembly probe is moved into position within a metallic guide channel. The guide channel penetrates the wall of the vessel and forms part of the barrier to the environment within the pressure vessel. Each channel includes a ball check valve which is opened by the instrument assembly probe when the probe passes through the valve. A ball valve element is moved from its seat by the probe to a position lateral of the bore of the channel and is guided to its seat along a sloped path within the valve body when the probe is removed. 5 claims, 3 figures

  3. Valve for gas centrifuges

    Science.gov (United States)

    Hahs, Charles A.; Burbage, Charles H.

    1984-01-01

    The invention is a pneumatically operated valve assembly for simultaneously (1) closing gas-transfer lines connected to a gas centrifuge or the like and (2) establishing a recycle path between two of the lines so closed. The valve assembly is especially designed to be compact, fast-acting, reliable, and comparatively inexpensive. It provides large reductions in capital costs for gas-centrifuge cascades.

  4. Coanda effect in valves

    Directory of Open Access Journals (Sweden)

    Uruba Václav

    2017-01-01

    Full Text Available Coanda effect takes place in flow within valves diffuser for certain conditions. The valve plug in half-closed position forms wall-jet, which could be stable or instable, depending on geometry and other conditions. This phenomenon was subject of experimental study using time-resolved PIV technique. For the acquired data analysis the special spatio-temporal methods have been used.

  5. Velocity measurements and flow patterns within the hinge region of a Medtronic Parallel bileaflet mechanical valve with clear housing.

    Science.gov (United States)

    Ellis, J T; Healy, T M; Fontaine, A A; Saxena, R; Yoganathan, A P

    1996-11-01

    . Although the LDA measurements were conducted at only a limited number of sites within the hinge, these results suggest that the hinge design can significantly affect the washout capacity and thrombogenic potential of the Medtronic Parallel bileaflet mechanical heart valve. The use of LDA within the confines of the hinge region of a mechanical heart valve is a new application, made possible by recent advances in manufacturing technologies and a proprietary process developed by Medtronic that allowed the production of a transparent valve housing. Together, these modalities represent a new method by which future valve designs can be assessed before clinical trials are initiated.

  6. Safety and relief valves in light water reactors

    International Nuclear Information System (INIS)

    Singh, A.

    1985-12-01

    Information is presented to: provide an introduction to and descriptions of various types of safety and relief valves in both PWR and BWR plants; describe anticipated operating conditions for these valves; describe the test facilities, procedures, and major results for both types of valves; present an extensive discussion of modeling and analysis of safety and relief valve performance, including the prediction of flow capacity and stability during operation; deal with the analyses related to the prediction of thermal-hydraulic loads on discharge piping and comparison against test data; discuss results of small-scale valve tests and flow visualization studies through transparent valve models; and describe an EPRI study for optimizing a typical PWR over-pressure protection system to enhance the availability and reliability of plant operation and thus reduce operation costs

  7. DESIGN OPTIMIZATION OF A FOOT VALVE BY USING ANSYS®

    Directory of Open Access Journals (Sweden)

    Serdar KARAOĞLU

    2008-02-01

    Full Text Available In this study, main components of a foot valve, being produced by casting, were optimized for minimum weight. The study was focused on the minimization of casting costs by reducing the volumes of two main parts of the foot valve. ANSYS® finite elements package was used in the study. In the optimization stage, parametrical dimensions were determined according to manufacturer's design criteria and related standards. Final design of the foot valve was completed by using the calculated values of optimum dimensions of the main components. Design optimization procedure gave about 8.5% of weight reductions in the main foot valve components.

  8. Improved hemocompatibility of silicone rubber extracorporeal tubing via solvent swelling-impregnation of S-nitroso-N-acetylpenicillamine (SNAP) and evaluation in rabbit thrombogenicity model.

    Science.gov (United States)

    Brisbois, Elizabeth J; Major, Terry C; Goudie, Marcus J; Bartlett, Robert H; Meyerhoff, Mark E; Handa, Hitesh

    2016-06-01

    Blood-contacting devices, including extracorporeal circulation (ECC) circuits, can suffer from complications due to platelet activation and thrombus formation. Development of nitric oxide (NO) releasing polymers is one method to improve hemocompatibility, taking advantage of the ability of low levels of NO to prevent platelet activation/adhesion. In this study a novel solvent swelling method is used to load the walls of silicone rubber tubing with the NO donor S-nitroso-N-acetylpenicillamine (SNAP). This SNAP-silicone rubber tubing exhibits an NO flux of ca. 1×10(-10)molcm(-2)min(-1), which mimics the range of NO release from the normal endothelium, which is stable for at least 4h. Images of the tubing before and after swelling, obtained via scanning electron microscopy, demonstrate that this swelling method has little effect on the surface properties of the tubing. The SNAP-loaded silicone rubber and silicone rubber control tubing are used to fabricate ECC circuits that are evaluated in a rabbit model of thrombogenicity. After 4h of blood flow, the SNAP-loaded silicone rubber circuits were able to preserve the blood platelet count at 64% of baseline (vs. 12% for silicone rubber control). A 67% reduction in the degree of thrombus formation within the thrombogenicity chamber was also observed. This study demonstrates the ability to improve the hemocompatibility of existing/commercial silicone rubber tubing via a simple solvent swelling-impregnation technique, which may also be applicable to other silicone-based blood-contacting devices. Localized nitric oxide (NO) release can be achieved from biomedical grade polymers doped with S-nitroso-N-acetylpenicillamine (SNAP). Despite the promising in vitro and in vivo biocompatibility results reported for these NO releasing polymers, many of these materials may face challenges in being translated to clinical applications, especially in the areas of polymer processing and manufacturing. In this study, we report a solvent

  9. Transcatheter Aortic Valve Replacement for Degenerative Bioprosthetic Surgical Valves

    DEFF Research Database (Denmark)

    Dvir, Danny; Webb, John; Brecker, Stephen

    2012-01-01

    Transcatheter aortic valve-in-valve implantation is an emerging therapeutic alternative for patients with a failed surgical bioprosthesis and may obviate the need for reoperation. We evaluated the clinical results of this technique using a large, worldwide registry....

  10. NRC valve performance test program - check valve testing

    International Nuclear Information System (INIS)

    Jeanmougin, N.M.

    1987-01-01

    The Valve Performance Test Program addresses the current requirements for testing of pressure isolation valves (PIVs) in light water reactors. Leak rate monitoring is the current method used by operating commercial power plants to survey the condition of their PIVs. ETEC testing of three check valves (4-inch, 6-inch, and 12-inch nominal diameters) indicates that leak rate testing is not a reliable method for detecting impending valve failure. Acoustic emission monitoring of check valves shows promise as a method of detecting loosened internals damage. Future efforts will focus on evaluation of acoustic emission monitoring as a technique for determining check valve condition. Three gate valves also will be tested to evaluate whether the check valve results are applicable to gate type PIVs

  11. Main feedwater valve diagnostics at Waterford 3 nuclear generating station

    International Nuclear Information System (INIS)

    Fitzgerald, W.V.

    1991-01-01

    Pneumatically-operated control valves are coming under increasing scrutiny in nuclear power plants because of their relatively high incident rate. The theory behind a device that could make performance evaluation of these valves simpler and more effective was first described at the original EPRI Power Plant Valve Symposium. The development of this Diagnostic System was completed in 1989, and it was recently used to troubleshoot two main feedwater valves at Louisiana Power and Light's Waterford 3 Power Station. During a cold snap last December, these valves failed to respond to the input signal and, as a result, the plant came off line. An incident report had to be filed, and the plant chose to contact the original equipment manufacturer (OEM) for assistance. This paper describes the original incident involving these valves and then gives a brief description of the diagnostic system and how it works. The balance of the paper then reviews how the OEM and plant personnel utilized the system to evaluate each component of the control valve assembly (I/P transducer, positioner, volume boosters, actuator, and valve body assembly). By simply stroking the valve and monitoring pneumatic signals and valve position, the problem was traced to a malfunctioning positioner and a volume booster that was leaking. The problems were corrected and new performance signatures run for the valves using the system to document their improved operation. This case study demonstrates how new Diagnostic Technology along with OEM involvement can effectively address problems with pneumatically-operated control valves so that root-cause solutions can be implemented

  12. Check valves aging assessment

    International Nuclear Information System (INIS)

    Haynes, H.D.

    1991-01-01

    In support of the NRC Nuclear Plant Aging Research (NPAR) program, the Oak Ridge National Laboratory (ORNL) has carried out an assessment of several check value diagnostic monitoring methods, in particular, those based on measurements of acoustic emission, ultrasonics, and magnetic flux. The evaluations have focussed on the capabilities of each method to provide information useful in determining check valve aging and service wear effects, check valve failures, and undesirable operating modes. This paper describes the benefits and limitations associated with each method and includes recent laboratory and field test data, including data obtained from the vendors who recently participated in a comprehensive series of tests directed by a nuclear industry users group. In addition, as part of the ORNL Advanced Diagnostic Engineering Research and Development Center (ADEC), two novel nonintrusive monitoring methods were developed that provide several unique capabilities. These methods, based on external ac- an dc-magnetic monitoring are also described. None of the examined methods could, by themselves, monitor both the instantaneous position and motion of check valve internals and valve leakage; however, the combination of acoustic emission monitoring with one of the other methods provides the means to determine vital check valve operational information

  13. Stop valve with automatic control and locking for nuclear reactors

    International Nuclear Information System (INIS)

    Chung, D.K.

    1980-01-01

    This invention generally concerns an automatic control and locking stop valve. Specifically it relates to the use of such a valve in a nuclear reactor of the type containing absorber elements supported by a fluid and intended for stopping the reactor in complete safety [fr

  14. Late complications in patients with Björk-Shiley and St. Jude Medical heart valve replacement.

    Science.gov (United States)

    Horstkotte, D; Körfer, R; Seipel, L; Bircks, W; Loogen, F

    1983-09-01

    Valve-related complications after Björk-Shiley mitral valve implantation (n = 475), aortic valve implantation (n = 424), or mitral-aortic valve implantation (n = 119) were compared with those after St. Jude Medical mitral valve replacement (n = 173), aortic valve replacement (n = 152), or mitral-aortic valve replacement (n = 69). All patients were placed on anticoagulant therapy with phenprocoumon early after operation. All patients had a comparable follow-up time of approximately 23 months, which showed that cumulative thromboembolic rates were significantly higher after St. Jude valve implantation than after Björk-Shiley valve implantation. Reoperations were necessary because of valve thrombosis (0.46%), perivalvular leakage (2.2%), or prosthetic valve endocarditis with perivalvular regurgitation (0.46%). One Björk-Shiley mitral valve prosthesis had to be replaced because of fracture of the outlet strut. Without significant intergroup differences, hemorrhage due to anticoagulant treatment was the most frequent complication. Thromboembolic complications were significantly more frequent after Björk-Shiley mitral, aortic, and double valve replacements than after St. Jude valve implantation. This may lead to consideration of changes in the prophylaxis of thrombus formations in the St. Jude valve, especially in aortic valve replacements, in patients with sinus rhythm.

  15. Comparative study of Butterfly valves

    International Nuclear Information System (INIS)

    Galmes Belmonte, F.B.

    1998-01-01

    This work tries to justify the hydrodynamic butterfly valves performance, using the EPRI tests, results carried out in laboratory and in situ. This justification will be possible if: - The valves to study are similar - Their performance is calculated using EPRI's methodology Looking for this objective, the elements of the present work are: 1. Brief EPRI butterfly valve description it wild provide the factors which are necessary to define the butterfly valves similarity. 2. EPRI tests description and range of validation against test data definition. 3. Description of the spanish butterfly analyzed valves, and comparison with the EPRI performance results, to prove that this valves are similar to the EPRI test valves. In this way, it will not be necessary to carry out particular dynamic tests on the spanish valves to describe their hydrodynamic performance. (Author)

  16. Coronary artery assessment by multidetector computed tomography in patients with prosthetic heart valves

    International Nuclear Information System (INIS)

    Habets, Jesse; Mali, Willem P.T.M.; Budde, Ricardo P.J.; Brink, Renee B.A. van den; Uijlings, Ruben; Spijkerboer, Anje M.; Chamuleau, Steven A.J.

    2012-01-01

    Patients with prosthetic heart valves may require assessment for coronary artery disease. We assessed whether valve artefacts hamper coronary artery assessment by multidetector CT. ECG-gated or -triggered CT angiograms were selected from our PACS archive based on the presence of prosthetic heart valves. The best systolic and diastolic axial reconstructions were selected for coronary assessment. Each present coronary segment was scored for the presence of valve-related artefacts prohibiting coronary artery assessment. Scoring was performed in consensus by two observers. Eighty-two CT angiograms were performed on a 64-slice (n = 27) or 256-slice (n = 55) multidetector CT. Eighty-nine valves and five annuloplasty rings were present. Forty-three out of 1160 (3.7%) present coronary artery segments were non-diagnostic due to valve artefacts (14/82 patients). Valve artefacts were located in right coronary artery (15/43; 35%), left anterior descending artery (2/43; 5%), circumflex artery (14/43; 32%) and marginal obtuse (12/43; 28%) segments. All cobalt-chrome containing valves caused artefacts prohibiting coronary assessment. Biological and titanium-containing valves did not cause artefacts except for three specific valve types. Most commonly implanted prosthetic heart valves do not hamper coronary assessment on multidetector CT. Cobalt-chrome containing prosthetic heart valves preclude complete coronary artery assessment because of severe valve artefacts. circle Most commonly implanted prosthetic heart valves do not hamper coronary artery assessment circle Prosthetic heart valve composition determines the occurrence of prosthetic heart valve-related artefacts circle Bjoerk-Shiley and Sorin tilting disc valves preclude diagnostic coronary artery segment assessment. (orig.)

  17. A modelling study into the effects of variable valve timing on the gas exchange process and performance of a 4-valve DI homogeneous charge compression ignition (HCCI) engine

    International Nuclear Information System (INIS)

    Mahrous, A-F.M.; Potrzebowski, A.; Wyszynski, M.L.; Xu, H.M.; Tsolakis, A.; Luszcz, P.

    2009-01-01

    Homogeneous charge compression ignition (HCCI) combustion mode is a relatively new combustion technology that can be achieved by using specially designed cams with reduced lift and duration. The auto-ignition in HCCI engine can be facilitated by adjusting the timing of the exhaust-valve-closing and, to some extent, the timing of the intake-valve-opening so as to capture a proportion of the hot exhaust gases in the engine cylinder during the gas exchange process. The effects of variable valve timing strategy on the gas exchange process and performance of a 4-valve direct injection HCCI engine were computationally investigated using a 1D fluid-dynamic engine cycle simulation code. A non-typical intake valve strategy was examined; whereby the intake valves were assumed to be independently actuated with the same valve-lift profile but at different timings. Using such an intake valves strategy, the obtained results showed that the operating range of the exhaust-valve-timing within which the HCCI combustion can be facilitated and maintained becomes much wider than that of the typical intake-valve-timing case. Also it was found that the engine parameters such as load and volumetric efficiency are significantly modified with the use of the non-typical intake-valve-timing. Additionally, the results demonstrated the potential of the non-typical intake-valve strategy in achieving and maintaining the HCCI combustion at much lower loads within a wide range of valve timings. Minimizing the pumping work penalty, and consequently improving the fuel economy, was shown as an advantage of using the non-typical intake-valve-timing with the timing of the early intake valve coupled with a symmetric degree of exhaust-valve-closing timing

  18. A modelling study into the effects of variable valve timing on the gas exchange process and performance of a 4-valve DI homogeneous charge compression ignition (HCCI) engine

    Energy Technology Data Exchange (ETDEWEB)

    Mahrous, A-F.M. [School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham B15 2TT (United Kingdom); Lecturer at the Department of Mechanical Power Engineering, Faculty of Engineering (Shebin El-Kom), Menoufiya University, Shebin El-Kom (Egypt); Potrzebowski, A.; Wyszynski, M.L.; Xu, H.M.; Tsolakis, A.; Luszcz, P. [School of Mechanical Engineering, University of Birmingham, Edgbaston, Birmingham B15 2TT (United Kingdom)

    2009-02-15

    Homogeneous charge compression ignition (HCCI) combustion mode is a relatively new combustion technology that can be achieved by using specially designed cams with reduced lift and duration. The auto-ignition in HCCI engine can be facilitated by adjusting the timing of the exhaust-valve-closing and, to some extent, the timing of the intake-valve-opening so as to capture a proportion of the hot exhaust gases in the engine cylinder during the gas exchange process. The effects of variable valve timing strategy on the gas exchange process and performance of a 4-valve direct injection HCCI engine were computationally investigated using a 1D fluid-dynamic engine cycle simulation code. A non-typical intake valve strategy was examined; whereby the intake valves were assumed to be independently actuated with the same valve-lift profile but at different timings. Using such an intake valves strategy, the obtained results showed that the operating range of the exhaust-valve-timing within which the HCCI combustion can be facilitated and maintained becomes much wider than that of the typical intake-valve-timing case. Also it was found that the engine parameters such as load and volumetric efficiency are significantly modified with the use of the non-typical intake-valve-timing. Additionally, the results demonstrated the potential of the non-typical intake-valve strategy in achieving and maintaining the HCCI combustion at much lower loads within a wide range of valve timings. Minimizing the pumping work penalty, and consequently improving the fuel economy, was shown as an advantage of using the non-typical intake-valve-timing with the timing of the early intake valve coupled with a symmetric degree of exhaust-valve-closing timing. (author)

  19. Nuclear reactor steam depressurization valve

    International Nuclear Information System (INIS)

    Moore, G.L.

    1991-01-01

    This patent describes improvement in a nuclear reactor plant, an improved steam depressurization valve positioned intermediate along a steam discharge pipe for controlling the venting of steam pressure from the reactor through the pipe. The improvement comprises: a housing including a domed cover forming a chamber and having a partition plate dividing the chamber into a fluid pressure activation compartment and a steam flow control compartment, the valve housing being provided with an inlet connection and an outlet connection in the steam flow control compartment, and a fluid duct in communication with a source of fluid pressure for operating the valve; a valve set mounted within the fluid flow control compartment comprising a cylindrical section surrounding the inlet connection with one end adjoining the connection and having a radially projecting flange at the other end with a contoured extended valve sealing flange provided with an annular valve sealing member, and a valve cylinder traversing the partition plate and reciprocally movable within an opening in the partition plate with one terminal and extending into the fluid pressure activation compartment and the other terminal end extending into the steam flow control compartment coaxially aligned with the valve seat surrounding the inlet connection, the valve cylinder being surrounded by two bellow fluid seals and provided with guides to inhibit lateral movement, an end of the valve cylinder extending into the fluid flow control compartment having a radially projecting flange substantially conterminous with the valve seat flange and having a contoured surface facing and complimentary to the contoured valve seating surface whereby the two contoured valve surfaces can meet in matching relationship, thus providing a pressure actuated reciprocatable valve member for making closing contact with the valve seat and withdrawing therefrom for opening fluid flow through the valve

  20. Pulmonary Valve Replacement : Twenty-Six Years of Experience With Mechanical Valvar Prostheses

    NARCIS (Netherlands)

    Freling, Hendrik G.; van Slooten, Ymkje J.; van Melle, Joost P.; Ebels, Tjark; Hoendermis, Elke S.; Berger, Rolf M. F.; Hillege, Hans L.; Waterbolk, Tjalling W.; van Veldhuisen, Dirk J.; Willems, Tineke P.; Pieper, Petronella G.

    BACKGROUND: Although the thromboembolic risk after pulmonary valve replacement (PVR) with mechanical valves is presumed to be high, recent studies suggest promising short-term and mid-term results. However, large studies reporting long-term mortality and valve-related complications are missing.

  1. Valve spindle gland

    International Nuclear Information System (INIS)

    Burda, Z.; Harazim, A.; Kerlin, K.

    1979-01-01

    A gland is proposed of the valve spindle designed for radioactive or otherwise harmful media, such as in nuclear power plant primary circuits. The gland is installed in the valve cover and consists of a primary and a secondary part and of a gland case partitioning the gland space into two chambers. The bottom face of the gland case is provided with a double-sided collar for controlling the elements of the bottom primary gland while the top face is provided with a removable flange. (M.S.)

  2. Building valve amplifiers

    CERN Document Server

    Jones, Morgan

    2013-01-01

    Building Valve Amplifiers is a unique hands-on guide for anyone working with tube audio equipment--as an electronics hobbyist, audiophile or audio engineer. This 2nd Edition builds on the success of the first with technology and technique revisions throughout and, significantly, a major new self-build project, worked through step-by-step, which puts into practice the principles and techniques introduced throughout the book. Particular attention has been paid to answering questions commonly asked by newcomers to the world of the valve, whether audio enthusiasts tackling their first build or

  3. Valve thrombosis following transcatheter aortic valve implantation: a systematic review.

    Science.gov (United States)

    Córdoba-Soriano, Juan G; Puri, Rishi; Amat-Santos, Ignacio; Ribeiro, Henrique B; Abdul-Jawad Altisent, Omar; del Trigo, María; Paradis, Jean-Michel; Dumont, Eric; Urena, Marina; Rodés-Cabau, Josep

    2015-03-01

    Despite the rapid global uptake of transcatheter aortic valve implantation, valve trombosis has yet to be systematically evaluated in this field. The aim of this study was to determine the clinical characteristics, diagnostic criteria, and treatment outcomes of patients diagnosed with valve thrombosis following transcatheter aortic valve implantation through a systematic review of published data. Literature published between 2002 and 2012 on valve thrombosis as a complication of transcatheter aortic valve implantation was identified through a systematic electronic search. A total of 11 publications were identified, describing 16 patients (mean age, 80 [5] years, 65% men). All but 1 patient (94%) received a balloon-expandable valve. All patients received dual antiplatelet therapy immediately following the procedure and continued to take either mono- or dual antiplatelet therapy at the time of valve thrombosis diagnosis. Valve thrombosis was diagnosed at a median of 6 months post-procedure, with progressive dyspnea being the most common symptom. A significant increase in transvalvular gradient (from 10 [4] to 40 [12] mmHg) was the most common echocardiographic feature, in addition to leaflet thickening. Thrombus was not directly visualized with echocardiography. Three patients underwent valve explantation, and the remaining received warfarin, which effectively restored the mean transvalvular gradient to baseline within 2 months. Systemic embolism was not a feature of valve thrombosis post-transcatheter aortic valve implantation. Although a rare, yet likely under-reported complication of post-transcatheter aortic valve implantation, progressive dyspnea coupled with an increasing transvalvular gradient on echocardiography within the months following the intervention likely signifies valve thrombosis. While direct thrombus visualization appears difficult, prompt initiation of oral anticoagulation therapy effectively restores baseline valve function. Copyright © 2014

  4. Cavitation problems in sodium valves

    International Nuclear Information System (INIS)

    Elie, X.

    1976-01-01

    Cavitation poses few problems for sodium valves, in spite of the fact that the loops are not pressurized. This is no doubt due to the low flow velocities in the pipes. For auxiliary loop valves we are attempting to standardize performances with respect to cavitation. For economic reasons cavitation thresholds are approached with large diameter valves. (author)

  5. Radiological findings of congenital urethral valves

    International Nuclear Information System (INIS)

    Yeon, Kyung Mo; Kook, Shin Ho

    1990-01-01

    Congenital urethral valve is the common cause of hydronephrosis in newborn infants and the most common cause of bladder outlet obstruction in male children. We reviewed and analysed radiological findings and associated anomalies of 16 cases of congenital urethral valve which were examined during the period from January 1985 to December 1989. The most frequent age was under one year old (56%). The main symptoms were urinary dribbing (37.5%), weak stream (25%) and urinary frequency and incontinence (25%). Anterior urethral valve (AUV) was 5 cases (31%) and posterior urethral valve (PUV) was 11 cases(69%), in which 10 cases were Type I and one case was Type III. Bladder wall thickening was seen in all cases and its severity was partly correlated with the degree of vesicoureteral reflux (VUR). VUR was observed in 12 cases (75%), and relatively severe in older age group. The degree of VUR was milder in AUV than PUV. Hydronephrosis was more severe in PUV than in anterior one, and its degree was correlated with the severity of VUR. Associated anomalies were ectopic urethral opening (2 cases), PDA (1 case), congenital megacolon (1 case) and patent urachus (1 case) in PUV. So early diagnosis and treatment of congenital urethral valve is essential to the prevention of renal damage

  6. Transcatheter aortic valve implantation in failed bioprosthetic surgical valves

    DEFF Research Database (Denmark)

    Dvir, Danny; Webb, John G; Bleiziffer, Sabine

    2014-01-01

    for patients with structural valve deterioration; however, a comprehensive evaluation of survival after the procedure has not yet been performed. OBJECTIVE: To determine the survival of patients after transcatheter valve-in-valve implantation inside failed surgical bioprosthetic valves. DESIGN, SETTING......, stroke, and New York Heart Association functional class. RESULTS: Modes of bioprosthesis failure were stenosis (n = 181 [39.4%]), regurgitation (n = 139 [30.3%]), and combined (n = 139 [30.3%]). The stenosis group had a higher percentage of small valves (37% vs 20.9% and 26.6% in the regurgitation...... and combined groups, respectively; P = .005). Within 1 month following valve-in-valve implantation, 35 (7.6%) patients died, 8 (1.7%) had major stroke, and 313 (92.6%) of surviving patients had good functional status (New York Heart Association class I/II). The overall 1-year Kaplan-Meier survival rate was 83...

  7. Fiber heart valve prosthesis: influence of the fabric construction parameters on the valve fatigue performances.

    Science.gov (United States)

    Vaesken, Antoine; Heim, Frederic; Chakfe, Nabil

    2014-12-01

    Transcatheter aortic valve replacement (TAVR) has become today a largely considered alternative technique to surgical valve replacement in patients who are not operable or patients with high risk for open chest surgery. However, the biological valve tissue used in the devices implanted clinically appears to be fragile material when folded for low diameter catheter insertion purpose and released in calcified environment with irregular geometry. Textile polyester material is characterized by outstanding folding and strength properties combined with proven biocompatibility. It could thereof be considered to replace biological valve leaflets in the TAVR procedure. The textile construction parameters must however be tuned to obtain a material compatible with the valve requested durability. In that context, one issue to be addressed is the friction effect that occurs between filaments and between yarns within a fabric under flexure loading. This phenomenon could be critical for the resistance of the material on the long term. The purpose of the present work is to assess the fatigue performances of textile valve prototypes made from different fabric constructions (monofilament, multifilament, calendered mutifilament) under accelerated cyclic loading. The goal is to identify, which construction is the best suited to long term fatigue stress. Results show that calendered multifilament and monofilament fabric constructions undergo strong ruptures already from 40 Mio cycles, while non calendered multifilament appears more durable. The rupture patterns observed point out that durability is directly related to the flexure stiffness level of the fibrous elements in the construction. Copyright © 2014 Elsevier Ltd. All rights reserved.

  8. Multimodality Imaging of Heart Valve Disease

    International Nuclear Information System (INIS)

    Rajani, Ronak; Khattar, Rajdeep; Chiribiri, Amedeo; Victor, Kelly; Chambers, John

    2014-01-01

    Unidentified heart valve disease is associated with a significant morbidity and mortality. It has therefore become important to accurately identify, assess and monitor patients with this condition in order that appropriate and timely intervention can occur. Although echocardiography has emerged as the predominant imaging modality for this purpose, recent advances in cardiac magnetic resonance and cardiac computed tomography indicate that they may have an important contribution to make. The current review describes the assessment of regurgitant and stenotic heart valves by multimodality imaging (echocardiography, cardiac computed tomography and cardiac magnetic resonance) and discusses their relative strengths and weaknesses

  9. Multimodality Imaging of Heart Valve Disease

    Energy Technology Data Exchange (ETDEWEB)

    Rajani, Ronak, E-mail: Dr.R.Rajani@gmail.com [Department of Cardiology, St. Thomas’ Hospital, London (United Kingdom); Khattar, Rajdeep [Department of Cardiology, Royal Brompton Hospital, London (United Kingdom); Chiribiri, Amedeo [Divisions of Imaging Sciences, The Rayne Institute, St. Thomas' Hospital, London (United Kingdom); Victor, Kelly; Chambers, John [Department of Cardiology, St. Thomas’ Hospital, London (United Kingdom)

    2014-09-15

    Unidentified heart valve disease is associated with a significant morbidity and mortality. It has therefore become important to accurately identify, assess and monitor patients with this condition in order that appropriate and timely intervention can occur. Although echocardiography has emerged as the predominant imaging modality for this purpose, recent advances in cardiac magnetic resonance and cardiac computed tomography indicate that they may have an important contribution to make. The current review describes the assessment of regurgitant and stenotic heart valves by multimodality imaging (echocardiography, cardiac computed tomography and cardiac magnetic resonance) and discusses their relative strengths and weaknesses.

  10. SAFETY SHUTOFF VALVE

    DEFF Research Database (Denmark)

    2010-01-01

    It is disclosed a shut-off valve which acts automatically and has a fully mechanical performance with respect to the loosing of the tower-shape part balance under the effect of the special acceleration Which is arisen from the quakes waves or serious vibrations, while such vibrations are mainly r...

  11. Thermostatic Radiator Valve Evaluation

    Energy Technology Data Exchange (ETDEWEB)

    Dentz, Jordan [Advanced Residential Integrated Energy Solutions Collaborative, New York, NY (United States); Ansanelli, Eric [Advanced Residential Integrated Energy Solutions Collaborative, New York, NY (United States)

    2015-01-01

    A large stock of multifamily buildings in the Northeast and Midwest are heated by steam distribution systems. Losses from these systems are typically high and a significant number of apartments are overheated much of the time. Thermostatically controlled radiator valves (TRVs) are one potential strategy to combat this problem, but have not been widely accepted by the residential retrofit market.

  12. Blocked Urethral Valves

    Science.gov (United States)

    ... if any damage has occurred to the upper urinary tract. Your pediatrician will consult with a pediatric nephrologist (kidney specialist) or nurologist, who may recommend surgery to remove the obstructing valves and prevent further infection or damage to the kidneys or urinary system. ...

  13. Methods of estimation of mitral valve regurgitation for the cardiac surgeon

    Directory of Open Access Journals (Sweden)

    Baikoussis Nikolaos G

    2009-07-01

    Full Text Available Abstract Mitral valve regurgitation is a relatively common and important heart valve lesion in clinical practice and adequate assessment is fundamental to decision on management, repair or replacement. Disease localised to the posterior mitral valve leaflet or focal involvement of the anterior mitral valve leaflet is most amenable to mitral valve repair, whereas patients with extensive involvement of the anterior leaflet or incomplete closure of the valve are more suitable for valve replacement. Echocardiography is the recognized investigation of choice for heart valve disease evaluation and assessment. However, the technique is depended on operator experience and on patient's hemodynamic profile, and may not always give optimal diagnostic views of mitral valve dysfunction. Cardiac catheterization is related to common complications of an interventional procedure and needs a hemodynamic laboratory. Cardiac magnetic resonance (MRI seems to be a useful tool which gives details about mitral valve anatomy, precise point of valve damage, as well as the quantity of regurgitation. Finally, despite of its higher cost, cardiac MRI using cine images with optimized spatial and temporal resolution can also resolve mitral valve leaflet structural motion, and can reliably estimate the grade of regurgitation.

  14. Dynamic testing of POSI-SEAL motor-operated butterfly valves using strain gages

    International Nuclear Information System (INIS)

    Richard, M.C.; Chiou, D.

    1994-01-01

    Utilities operating nuclear power plants recognize that the correct functioning of all motor-operated valves, and particularly those in safety-related systems, is of paramount importance. The U.S. Nuclear Regulatory Commission has issued Generic Letter 89-10 relative to this concern. Operability must be demonstrated under design-basis conditions. In order to demonstrate operability of motor-operated butterfly valves, the valve stem torque must be determined. The valve stem torque is a function of seat material, stem packing, stem bearing friction, and hydrodynamic lift and drag. The total valve operating hydrodynamic torque can be predicted using the valve manufacturer's data and the differential pressure. In order to validate the valve manufacturer's data, the actual total valve hydrodynamic torque is measured using strain gages mounted directly on the valve stem. This paper presents the results of comparing the predicted total valve operating hydrodynamic torque with the actual total valve operating hydrodynamic torque for six POSI-SEAL Class 150 high performance butterfly valves

  15. Diaphragms in air-operated valves

    Energy Technology Data Exchange (ETDEWEB)

    Groeger, J.E.

    1996-12-01

    The author will present current issues related to diaphgrams in air-operated valves. Altran Materials Engineering, Inc., often performs root-cause analyses for nuclear power plant owners. The author will discuss various analyses that have been performed or are currently underway.

  16. Autopsy after transcatheter aortic valve implantation

    NARCIS (Netherlands)

    van Kesteren, F.; Wiegerinck, E. M. A.; Rizzo, S.; Baan, J.; Planken, R. N.; von der Thüsen, J. H.; Niessen, H. W. M.; van Oosterhout, M. F. M.; Pucci, A.; Thiene, G.; Basso, C.; Sheppard, M. N.; Wassilew, K.; van der Wal, A. C.

    2017-01-01

    Autopsy after transcatheter aortic valve implantation (TAVI) is a new field of interest in cardiovascular pathology. To identify the cause of death, it is important to be familiar with specific findings related to the time interval between the procedure and death. We aimed to provide an overview of

  17. Impact of Chronic Rheumatic Valve Diseases on Large Vessels.

    Science.gov (United States)

    Altunbas, Gokhan; Yuce, Murat; Ozer, Hasan O; Davutoglu, Vedat; Ercan, Suleyman; Kizilkan, Nese; Bilici, Muhammet

    2016-01-01

    BACKGROUND AND AIM OF STUDY: Rheumatic valvular heart disease, which remains a common health problem in developing countries, has numerous consequences on the heart chambers and circulation. The study aim was to investigate the effects of chronic rheumatic valve disease on the diameters of the descending aorta (DA) and inferior vena cava (IVC). METHODS: A total of 88 patients with echocardiographically documented rheumatic valvular heart disease and 112 healthy controls were enrolled into the study. All patients underwent detailed echocardiographic examinations, while their height and body weight were recorded and adjusted to their body surface area. RESULTS: The most common involvement was mitral valve disease, followed by aortic valve disease and tricuspid valve disease. The mean diameter of the DA (indexed to BSA) was 1.79 ± 0.49 cm for patients and 1.53 ± 0.41 for controls (p Rheumatic valve disease, especially mitral stenosis, was closely related to remodeling of the great vessels.

  18. Aortic valve ochronosis: a rare manifestation of alkaptonuria.

    Science.gov (United States)

    Steger, Christina Maria

    2011-07-28

    Alkaptonuric ochronosis is a heritable disorder of tyrosine metabolism, with various systemic abnormalities related to pigment deposition and degeneration of collagen and other tissues, including the heart and aorta. A 65-year-old woman with alkaptonuric ochronosis and a history of four joint replacements required aortic valve replacement for severe aortic stenosis. Operative findings included ochronosis of a partly calcified aortic valve and the aortic intima. The aortic valve was removed at surgery and histologically investigated. Light microscopic examination of the aortic valve revealed intracellular and extracellular deposits of ochronotic pigment and a chronic inflammatory infiltrate. Beside the case representation, the disease history, aetiology, pathogenesis, clinical presentation and treatment of aortic valve ochronosis are reviewed.

  19. Midterm Results of Aortic Valve Replacement with Cryopreserved Homografts

    Directory of Open Access Journals (Sweden)

    Emre Özker

    2012-06-01

    Full Text Available Objective: The aim of this study was to analyze the midterm clinical results of aortic valve replacement with cryopreserved homografts.Materials and Methods: Aortic valve replacement was performed in 40 patients with cryopreserved homograft. The indications were aortic valve endocarditis in 20 patients (50%, truncus arteriosus in 6 patients (15%, and re-stenosis or regurtitation after aortic valve reconstruction in 14 (35% patients. The valve sizes ranged from 10 to 27mm. A full root replacement technique was used for homograft replacement in all patients.Results: The 30-day postoperative mortality rate was 12.5% (5 patients. There were four late deaths. Only one of them was related to cardiac events. Overall mortality was 22.5%. Thirty-three patients were followed up for 67±26 months. Two patients needed reoperation due to aortic aneurysm caused by endocarditis. The mean transvalvular gradient significantly decreased after valve replacement (p<0.003. The last follow up showed that the 27 (82% patients had a normal left ventricular function.Conclusion: Cryopreserved homografts are safe alternatives to mechanical valves that can be used when there are proper indications. Although it has a high perioperative mortality rate, cryopreserved homograft implantation is an alternative for valve replacement, particularly in younger patients and for complex surgical problems such as endocarditis that must be minimalized.

  20. Extracellular Matrix Biomarker, Fibulin-1, Is Closely Related to NT-proBNP and Soluble Urokinase Plasminogen Activator Receptor in Patients with Aortic Valve Stenosis (The SEAS Study)

    DEFF Research Database (Denmark)

    Kruger, Ruan; Rasmussen, Lars M; Argraves, William S

    2014-01-01

    associated with subclinical atherosclerosis. Therefore, we aimed to explore the interplay between these biomarkers and mild to moderate aortic valve stenosis (AS). METHODS: In 374 patients with mild to moderate AS, we investigated the relationship of fibulin-1 with NT-proBNP, levels of suPAR and the degree.......01), and suPAR (βyear0 = 0.09, p = 0.26, βyear1 = 0.23, βyear4 = 0.21, both plevels of fibulin-1 were independently associated with higher levels of suPAR and NT-proBNP especially in patients with lower AVAI, suggesting...

  1. Minimising life cycle costs of automated valves in offshore platforms

    Energy Technology Data Exchange (ETDEWEB)

    Yli-Petays, Juha [Metso Automation do Brasil Ltda., Rio de Janeiro, RJ (Brazil); Niemela, Ismo [Metso Automation, Imatra (Finland)

    2012-07-01

    Automated process valves play an essential role in offshore platforms operation. If you are able to optimize their operation and maintenance activities you can receive extensive operational savings with minimal investment. Valves used in offshore platforms doesn't differentiate that much from the valves used in downstream but there are certain specialties, which makes the operations more challenging in offshore: Process valves are more difficult to access and maintain because of space limitations. Also spare part inventories and deliveries are challenging because of offshore platform's remote location. To overcome these challenges usage of digital positioners with diagnostic features has become more common because predictive maintenance capabilities enable possibilities to plan the maintenance activities and this way optimise the spare part orders regarding to valves. There are intelligent controllers available for control valves, automated on/off valves as well as ESD-valves and whole network of automated valves on platforms can be controlled by intelligent valve controllers. This creates many new opportunities in regards of optimized process performance or predictive maintenance point-of-view. By means of intelligent valve controllers and predictive diagnostics, condition monitoring and maintenance planning can also be performed remotely from an onshore location. Thus, intelligent valve controllers provide good way to minimize spending related to total cost of ownership of automated process valves. When purchase value of control valve represent 20% of TCO, intelligent positioner and predictive maintenance methods can enable as high as 30% savings over the life cycle of asset so basically it benefit savings higher than whole investment of monitored asset over its life cycle. This is mainly achieved through the optimized maintenance activities since real life examples has shown that with time based maintenance (preventive maintenance) approach 70% of

  2. Cyclonic valve test: preliminary results

    Energy Technology Data Exchange (ETDEWEB)

    Monteiro, Andre Sampaio; Moraes, Carlos Alberto C.; Marins, Luiz Philipe M.; Soares, Fabricio; Oliveira, Dennis; Lima, Fabio Soares de; Airao, Vinicius [Petroleo Brasileiro S.A. (PETROBRAS), Rio de Janeiro, RJ (Brazil); Ton, Tijmen [Twister BV, Rijswijk (Netherlands)

    2012-07-01

    For many years, the petroleum industry has been developing a valve that input less shear to the flow for a given required pressure drop and this can be done using the cyclonic concept. This paper presents a comparison between the performances of a cyclonic valve (low shear) and a conventional globe valve. The aim of this work is to show the advantages of using a cyclonic low shear valve instead of the commonly used in the primary separation process by PETROBRAS. Tests were performed at PETROBRAS Experimental Center (NUEX) in Aracaju/SE varying some parameters: water cut; pressure loss (from 4 kgf/cm2 to 10 kgf/cm2); flow rates (30 m3/h and 45 m3/h). Results indicates a better performance of the cyclonic valve, if compared with a conventional one, and also that the difference of the performance, is a function of several parameters (emulsion stability, water content free, and oil properties). The cyclonic valve tested can be applied as a choke valve, as a valve between separation stages (for pressure drop), or for controlling the level of vessels. We must emphasize the importance to avoid the high shear imposed by conventional valves, because once the emulsion is created, it becomes more difficult to break it. New tests are being planned to occur in 2012, but PETROBRAS is also analyzing real cases where the applications could increase the primary process efficiency. In the same way, the future installations are also being designed considering the cyclonic valve usage. (author)

  3. Valve which can be remotely handled and operated, particularly for the hot area of radioactive plants

    International Nuclear Information System (INIS)

    Radovan, G.; Sandling, M.J.; Davidson, J.W.; Blaseck, K.; Hoffmeister, L.; Westendorf, H.

    1986-01-01

    The valve is built up so that one replacement part contains all the wear parts of the valve, while the non-wearing parts and particularly the valve housing are solidly connected to the pipeline. The replacement part and the valve housing are fixed on alternate sides by a clamp. The replacement part has a central screw which can be moved relative to the other components, which is supported in both axial directions in a clamp which can be opened. The replacement part can thus be pressed into the valve housing and pulled out of it. (orig./HP) [de

  4. Floppy Mitral Valve (FMV) - Mitral Valve Prolapse (MVP) - Mitral Valvular Regurgitation and FMV/MVP Syndrome.

    Science.gov (United States)

    Boudoulas, Konstantinos Dean; Pitsis, Antonios A; Boudoulas, Harisios

    2016-01-01

    Mitral valve prolapse (MVP) results from the systolic movement of a portion(s) or segment(s) of the mitral valve leaflet(s) into the left atrium during left ventricular (LV) systole. It should be emphasised that MVP alone, as defined by imaging techniques, may comprise a non-specific finding because it also depends on the LV volume, myocardial contractility and other LV hemodynamics. Thus, a floppy mitral valve (FMV) should be the basis for the diagnosis of MVP. Two types of symptoms may be defined in these patients. In one group, symptoms are directly related to progressive mitral regurgitation and its complications. In the other group, symptoms cannot be explained only by the degree of mitral regurgitation alone; neuroendocrine dysfunction has been implicated for the explanation of symptoms in this group of patients that today is referred as the FMV/MVP syndrome. When significant mitral regurgitation is present in a patient with FMV/MVP, surgical intervention is recommended. In patients with a prohibitive risk for surgery, transcatheter mitral valve repair using a mitraclip device may be considered. Furthermore, transcatheter mitral valve replacement may represent an option in the near future as clinical trials are underway. In this brief review, the current concepts related to FMV/MVP and FMV/MVP syndrome will be discussed. Copyright © 2016 Hellenic Cardiological Society. Published by Elsevier B.V. All rights reserved.

  5. Posterior Urethral Valves

    Directory of Open Access Journals (Sweden)

    Steve J. Hodges

    2009-01-01

    Full Text Available The most common cause of lower urinary tract obstruction in male infants is posterior urethral valves. Although the incidence has remained stable, the neonatal mortality for this disorder has improved due to early diagnosis and intensive neonatal care, thanks in part to the widespread use of prenatal ultrasound evaluations. In fact, the most common reason for the diagnosis of posterior urethral valves presently is the evaluation of infants for prenatal hydronephrosis. Since these children are often diagnosed early, the urethral obstruction can be alleviated rapidly through catheter insertion and eventual surgery, and their metabolic derangements can be normalized without delay, avoiding preventable infant mortality. Of the children that survive, however, early diagnosis has not had much effect on their long-term prognosis, as 30% still develop renal insufficiency before adolescence. A better understanding of the exact cause of the congenital obstruction of the male posterior urethra, prevention of postnatal bladder and renal injury, and the development of safe methods to treat urethral obstruction prenatally (and thereby avoiding the bladder and renal damage due to obstructive uropathy are the goals for the care of children with posterior urethral valves[1].

  6. Anterior Urethral Valves

    Directory of Open Access Journals (Sweden)

    Vidyadhar P. Mali

    2006-07-01

    Full Text Available We studied the clinical presentation and management of four patients with anterior urethral valves; a rare cause of urethral obstruction in male children. One patient presented antenatally with oligohydramnios, bilateral hydronephrosis and bladder thickening suggestive of an infravesical obstruction. Two other patients presented postnatally at 1 and 2 years of age, respectively, with poor stream of urine since birth. The fourth patient presented at 9 years with frequency and dysuria. Diagnosis was established on either micturating cystourethrogram (MCU (in 2 or on cystoscopy (in 2. All patients had cystoscopic ablation of the valves. One patient developed a postablation stricture that was resected with an end-to-end urethroplasty. He had an associated bilateral vesicoureteric junction (VUJ obstruction for which a bilateral ureteric reimplantation was done at the same time. On long-term follow-up, all patients demonstrated a good stream of urine. The renal function is normal. Patients are continent and free of urinary infections. Anterior urethral valves are rare obstructive lesions in male children. The degree of obstruction is variable, and so they may present with mild micturition difficulty or severe obstruction with hydroureteronephrosis and renal impairment. Hence, it is important to evaluate the anterior urethra in any male child with suspected infravesical obstruction. The diagnosis is established by MCU or cystoscopy and the treatment is always surgical, either a transurethral ablation or an open resection. The long-term prognosis is good.

  7. High Reliability Cryogenic Piezoelectric Valve Actuator, Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — Piezoelectric actuators constructed with the "smart material" PZT offer many potential advantages for use in NASA cryo-valve missions relative to conventional...

  8. Platelet aggregation, eicosanoid production and thrombogenic ratio in individuals at high cardiovascular risk consuming meat enriched in walnut paste. A crossover, placebo-controlled study.

    Science.gov (United States)

    Canales, Amaia; Bastida, Sara; Librelottto, Josana; Nus, Meritxell; Sánchez-Muniz, Francisco J; Benedi, Juana

    2009-07-01

    Walnut consumption produces beneficial cardiovascular effects. The aim of the present study is to compare the effects of meat enriched in walnut paste (WM) and low-fat meat (LM) consumptions on platelet aggregation, plasma thromboxane A2 (TXA2, measured as TXB2), prostacyclin I2 (PGI2, as 6-keto-PGF1alpha) and the thrombogenic ratio (TXB2/6-keto-PGF1alpha) in volunteers at high CVD risk. Twenty-two adults were placed on a random, non-blinded crossover study involving two test periods (five portions WM/week for 5 week; five portions LM/week for 5 week) separated by a 4- to 6-week washout period. The participants were asked to complete a diet record throughout the study. Platelet aggregation, plasma TXB2, 6-keto-PGF1alpha production and the TXB2/6-keto-PGF1alpha ratio were determined at baseline and at weeks 3 and 5 for the two dietary periods. The WM diet contains a lower SFA content, a higher concentration of PUFA and a more favourable n-6/n-3 ratio than the LM diet. Significant time x treatment interactions were observed for TXB2 (P = 0.048) and the TXB2/6-keto-PGF1alpha ratio (P = 0.028). The WM diet significantly increased the level of 6-keto-PGF1alpha (P = 0.037) and decreased the TXB2/6-keto-PGF1alpha ratio (P = 0.048). At week 5, significant differences (P keto-PGF1alpha ratio. The effects on TXB2 and the TXB2/6-keto-PGF1alpha ratio were time-course dependent (P = 0.019 and 0.011, respectively). The WM and LM diets reduced TXB2 levels most (P = 0.050) in obese individuals, while the TXB2/6-keto-PGF1alpha ratio decreased most (P = 0.066) in volunteers whose serum cholesterol levels were > or = 2200 mg/l. The WM diet should be considered a functional meat because it improves the thrombogenic status mainly in individuals with high-cholesterol levels or high BMI.

  9. Valve-sparing and valve-replacing techniques for aortic root replacement in patients with Marfan syndrome: Analysis of early outcome.

    Science.gov (United States)

    Volguina, Irina V; Miller, D Craig; LeMaire, Scott A; Palmero, Laura C; Wang, Xing Li; Connolly, Heidi M; Sundt, Thoralf M; Bavaria, Joseph E; Dietz, Harry C; Milewicz, Dianna M; Coselli, Joseph S

    2009-05-01

    A prospective, international registry study was initiated to provide contemporary comparative data on short-term clinical outcomes after aortic valve-sparing and aortic valve-replacing root operations in patients with Marfan syndrome. The purpose of this initial report is to describe the study design and to compare early outcomes in the first 151 enrolled patients. We assessed 30-day outcomes in 151 patients who met strict Ghent diagnostic criteria for Marfan syndrome and underwent aortic root replacement with either valve-replacing (n = 46) or valve-sparing techniques (n = 105) at one of 18 participating centers. In the valve replacement group, a mechanical composite valve graft was used in 39 (85%) patients and a bioprosthetic valve in 7 (15%). In the valve-sparing group, David V procedures were performed in 57 (54%) patients, David I in 38 (36%), David IV in 8 (8%), Florida sleeve in 1 (1%), and Yacoub remodeling in 1 (1%). No in-hospital or 30-day deaths occurred. Despite longer crossclamp and cardiopulmonary bypass times in the valve-sparing group, there were no significant between-group differences in postoperative complications. Thirty-day valve-related complications occurred in 2 (4%) patients undergoing valve replacement and in 3 (3%) undergoing valve-sparing procedures (P = .6). The analysis of early outcomes revealed that valve-sparing techniques were the most common approach to root replacement in patients with Marfan syndrome in these centers. The complexity of valve-sparing root replacement did not translate into any demonstrable adverse early outcomes. Subsequent analysis will compare the 3-year durability of these two surgical approaches.

  10. Fracturing mechanics before valve-in-valve therapy of small aortic bioprosthetic heart valves.

    Science.gov (United States)

    Johansen, Peter; Engholt, Henrik; Tang, Mariann; Nybo, Rasmus F; Rasmussen, Per D; Nielsen-Kudsk, Jens Erik

    2017-10-13

    Patients with degraded bioprosthetic heart valves (BHV) who are not candidates for valve replacement may benefit from transcatheter valve-in-valve (VIV) therapy. However, in smaller-sized surgical BHV the resultant orifice may become too narrow. To overcome this, the valve frame can be fractured by a high-pressure balloon prior to VIV. However, knowledge on fracture pressures and mechanics are prerequisites. The aim of this study was to identify the fracture pressures needed in BHV, and to describe the fracture mechanics. Commonly used BHV of small sizes were mounted on a high-pressure balloon situated in a biplane fluoroscopic system with a high-speed camera. The instant of fracture was captured along with the balloon pressure. The valves were inspected for material protrusion and later dissected for fracture zone investigation and description. The valves with a polymer frame fractured at a lower pressure (8-10 atm) than those with a metal stent (19-26 atm). None of the fractured valves had elements protruding. VIV procedures in small-sized BHV may be performed after prior fracture of the valve frame by high-pressure balloon dilatation. This study provides tentative guidelines for expected balloon sizes and pressures for valve fracturing.

  11. Proximal thoracic aorta dimensions after continuous-flow left ventricular assist device implantation: Longitudinal changes and relation to aortic valve insufficiency.

    Science.gov (United States)

    Fine, Nowell M; Park, Soon J; Stulak, John M; Topilsky, Yan; Daly, Richard C; Joyce, Lyle D; Pereira, Naveen L; Schirger, John A; Edwards, Brooks S; Lin, Grace; Kushwaha, Sudhir S

    2016-04-01

    In this study we examined the impact of continuous-flow left ventricular assist device (CF-LVAD) support on proximal thoracic aorta dimensions. Aortic root and ascending aorta diameter were measured from serial echocardiograms before and after CF-LVAD implantation in patients with ≥6 months of support, and correlated with the development of >mild aortic valve insufficiency (AI). Of 162 patients included, mean age was 58 ± 11 years and 128 (79%) were male. Seventy-nine (63%) were destination therapy patients. Mean aortic root and ascending aorta diameters at baseline, 1 month, 6 months, 12 months and long-term follow-up (mean 2.0 ± 1.4 years) were 3.5 ± 0.4, 3.5 ± 0.3, 3.9 ± 0.3, 3.9 ± 0.2 and 4.0 ± 0.3, and 3.3 ± 0.2, 3.3 ± 0.3, 3.6 ± 0.2, 3.6 ± 0.3 and 3.6 ± 0.3 cm, respectively. Only change in aortic root diameter from 1-month to 6-month follow-up reached statistical significance (p = 0.03). Nine (6%) patients had accelerated proximal thoracic aorta expansion (>0.5 cm/year), occurring predominantly in the first 6 months after implantation. These patients were older and more likely to have hypertension and baseline proximal thoracic aorta dilation. Forty-five (28%) patients developed >mild AI at long-term follow-up, including 7 of 9 (78%) of those with accelerated proximal thoracic aorta expansion. All 7 had aortic valves that remained closed throughout the cardiac cycle, and this, along with duration of CF-LVAD support and increase in aortic root diameter, were significantly associated with developing >mild AI. CF-LVAD patients have small increases in proximal thoracic aorta dimensions that predominantly occur within the first 6 months after implantation and then stabilize. Increasing aortic root diameter was associated with AI development. Copyright © 2016 International Society for Heart and Lung Transplantation. Published by Elsevier Inc. All rights reserved.

  12. Automatic fire hydrant valve development

    International Nuclear Information System (INIS)

    Drumheller, K.

    1976-01-01

    The development of a remotely-controlled valve to operate a fire hydrant is described. Assembled from off-the-shelf components, the prototype illustrates that a valve light enough to be handled by one man is possible. However, it does not have the ruggedness or reliability needed for actual fire-fighting operations. Preliminary testing by City of Tacoma fire department personnel indicates that the valve may indeed contribute significantly to fire-fighting efficiency

  13. [Ahmed valve in glaucoma surgery].

    Science.gov (United States)

    Bikbov, M M; Khusnitdinov, I I

    This is a review on Ahmed valve application in glaucoma surgery. It contains, in particular, data on the Ahmed valve efficiency, results of experimental and histological studies of filtering bleb encapsulation, examines the use of antimetabolites and anti-VEGF agents, and discusses implantation techniques. The current appraisal of antimetabolites delivery systems integrated into the Ahmed valve is presented. Various complications encountered in practice and preventive measures are also covered.

  14. BIF butterfly valve life extension at WNP-2

    International Nuclear Information System (INIS)

    Armstrong, D.

    1991-01-01

    Primary containment purging, venting, inerting, and reactor building ventilation at the WNP-2 plant are accomplished with a series of large butterfly valves. A total of 31 valves which are similar in design, but of different sizes, employ an elastomer to achieve sealing integrity when closed. These valves, which were originally manufactured by BIF, a unit of General Signal, range in size from 18 to 84 inches in diameter. Service life in the plant was much less than desired for safety-related equipment, and several seal failures had been experienced shortly after valve overhaul. This program covers a design change made to enhance performance of the elastomer seal to achieve a very meaningful life extension. While numerous configurations of BIF valves exist, this work relates only to the model 657 unit assembled with an elastomer seal mounted onto the valve disc by a stainless steel clamping ring held with studs and nuts. The problems encountered, and the steps taken to resolve the deficiencies may, however, be applicable to other butterfly valve configurations

  15. Thermal fatigue behavior of valves

    International Nuclear Information System (INIS)

    Moinereau, D.; Scliffet, L.; Capion, J.C.; Genette, P.

    1991-01-01

    This paper reports that valves of pressurized water reactors are exposed to thermal shocks during transient operations. The numerous thermal shock tests performed on valves on the EDF test facilities have shown the sensibility of fillets and geometrical discontinuities to thermal fatigue: cracks can appear in those areas and grow through the valve body. Valves systems designated as level 1 must be designed to withstand fatigue up to the second isolation valve: the relevant rule is specified in the paragraph B 3500 of the French RCCM code. It is a simplified method which doesn't require finite element calculations. Many valve systems have been designed according to this rule and have been operated without accident. However, in one case, important cracks were found in the fillet of a check-valve after numerous thermal shocks. Calculation of the valve's behavior according to the RCCM code to estimate the fatigue damage resulting from thermal shocks led to a low damage factor, which doesn't agree with the experimental results. This was confirmed by new testings and showed the inadequacy of B 3500 rule for thermal transients. On this base a new rule is proposed to estimate fatigue damage resulting from thermal shocks. An experimental program has been realized to validate this rule. Axisymetrical analytical mock-ups with different geometries and one check-valve in austenitic stainless steel 316 L have been submitted to hot thermal shocks of 210 degrees C magnitude

  16. Latest design of gate valves

    Energy Technology Data Exchange (ETDEWEB)

    Kurzhofer, U.; Stolte, J.; Weyand, M.

    1996-12-01

    Babcock Sempell, one of the most important valve manufacturers in Europe, has delivered valves for the nuclear power industry since the beginning of the peaceful application of nuclear power in the 1960s. The latest innovation by Babcock Sempell is a gate valve that meets all recent technical requirements of the nuclear power technology. At the moment in the United States, Germany, Sweden, and many other countries, motor-operated gate and globe valves are judged very critically. Besides the absolute control of the so-called {open_quotes}trip failure,{close_quotes} the integrity of all valve parts submitted to operational forces must be maintained. In case of failure of the limit and torque switches, all valve designs have been tested with respect to the quality of guidance of the gate. The guidances (i.e., guides) shall avoid a tilting of the gate during the closing procedure. The gate valve newly designed by Babcock Sempell fulfills all these characteristic criteria. In addition, the valve has cobalt-free seat hardfacing, the suitability of which has been proven by friction tests as well as full-scale blowdown tests at the GAP of Siemens in Karlstein, West Germany. Babcock Sempell was to deliver more than 30 gate valves of this type for 5 Swedish nuclear power stations by autumn 1995. In the presentation, the author will report on the testing performed, qualifications, and sizing criteria which led to the new technical design.

  17. Surge-damping vacuum valve

    International Nuclear Information System (INIS)

    Bullock, J.C.; Kelley, B.E.

    1977-01-01

    A valve for damping out flow surges in a vacuum system is described. The surge-damping mechanism consists of a slotted, spring-loaded disk adjacent to the valve's vacuum port (the flow passage to the vacuum roughing pump). Under flow surge conditions, the differential pressure forces the disk into a sealing engagement with the vacuum port, thereby restricting the gas flow path to narrow slots in the disk's periphery. The increased flow damps out the flow surge. When pressure is equalized on both sides of the valve, the spring load moves the disk away from the port to restore full flow conductance through the valve

  18. Interlayer quality dependent graphene spin valve

    International Nuclear Information System (INIS)

    Iqbal, Muhammad Zahir; Hussain, Ghulam; Siddique, Salma; Iqbal, Muhammad Waqas; Murtaza, Ghulam; Ramay, Shahid Mahmood

    2017-01-01

    It is possible to utilize the new class of materials for emerging two-dimensional (2D) spintronic applications. Here, the role of defects in the graphene interlayer and its influence on the spin valve signal is reported. The emergence of D peak in Raman spectrum reveals defects in the graphene layer. The linear I-V curve for defective and non-defective graphene samples indicate the ohmic nature of NiFe and graphene contact. A non-uniform magnetoresistive effect with a bump is persistently observed for defective graphene device at various temperatures, while a smooth and symmetric signal is detected for non-defective graphene spin valve. Parallel and antiparallel alignments of magnetization of magnetic materials shows low and high resistance states, respectively. The magnetoresistance (MR) ratio for defective graphene NiFe/graphene/NiFe spin valve is measured to be ~0.16% at 300 K which progresses to ~0.39% for non-defective graphene device at the same temperature. Similarly at 4.2 K the MR ratios are reported to be ~0.41% and ~0.78% for defective and non-defective graphene devices, respectively. Our investigation provides an evidence for relatively better response of the spin valve signal with high quality graphene interlayer.

  19. Graphene spin valve: An angle sensor

    Energy Technology Data Exchange (ETDEWEB)

    Iqbal, Muhammad Zahir, E-mail: zahir.upc@gmail.com [Faculty of Engineering Sciences, GIK Institute of Engineering Sciences and Technology, Topi 23640, Khyber Pakhtunkhwa (Pakistan); Hussain, Ghulam [Faculty of Engineering Sciences, GIK Institute of Engineering Sciences and Technology, Topi 23640, Khyber Pakhtunkhwa (Pakistan); Siddique, Salma [Department of Bioscience & Biotechnology, Sejong University, Seoul 143-747 (Korea, Republic of); Iqbal, Muhammad Waqas [Department of Physics, Riphah Institute of Computing and Applied Sciences (RICAS), Riphah International University, Lahore (Pakistan)

    2017-06-15

    Graphene spin valves can be optimized for various spintronic applications by tuning the associated experimental parameters. In this work, we report the angle dependent magnetoresistance (MR) in graphene spin valve for different orientations of applied magnetic field (B). The switching points of spin valve signals show a clear shift towards higher B for each increasing angle of the applied field, thus sensing the response for respective orientation of the magnetic field. The angular variation of B shifts the switching points from ±95 G to ±925 G as the angle is varied from 0° to 90° at 300 K. The observed shifts in switching points become more pronounced (±165 G to ±1450 G) at 4.2 K for similar orientation. A monotonic increase in MR ratio is observed as the angle of magnetic field is varied in the vertical direction at 300 K and 4.2 K temperatures. This variation of B (from 0° to 90°) increases the magnitude of MR ratio from ∼0.08% to ∼0.14% at 300 K, while at 4.2 K it progresses to ∼0.39% from ∼0.14%. The sensitivity related to angular variation of such spin valve structure can be employed for angle sensing applications.

  20. Interlayer quality dependent graphene spin valve

    Energy Technology Data Exchange (ETDEWEB)

    Iqbal, Muhammad Zahir, E-mail: zahir.upc@gmail.com [Faculty of Engineering Sciences, GIK Institute of Engineering Sciences and Technology, Topi, Khyber Pakhtunkhwa, 23640 Pakistan (Pakistan); Hussain, Ghulam [Faculty of Engineering Sciences, GIK Institute of Engineering Sciences and Technology, Topi, Khyber Pakhtunkhwa, 23640 Pakistan (Pakistan); Siddique, Salma [Department of Bioscience & Biotechnology, Sejong University, Seoul, 143-747 (Korea, Republic of); Iqbal, Muhammad Waqas [Department of Physics, Riphah Institute of Computing and Applied Sciences (RICAS), Riphah International University, Lahore (Pakistan); Murtaza, Ghulam [Centre for Advanced Studies in Physics, Government College University, Lahore 54000 (Pakistan); Ramay, Shahid Mahmood [Physics & Astronomy Department, College of Science, King Saud University, Riyadh 11451 (Saudi Arabia)

    2017-01-15

    It is possible to utilize the new class of materials for emerging two-dimensional (2D) spintronic applications. Here, the role of defects in the graphene interlayer and its influence on the spin valve signal is reported. The emergence of D peak in Raman spectrum reveals defects in the graphene layer. The linear I-V curve for defective and non-defective graphene samples indicate the ohmic nature of NiFe and graphene contact. A non-uniform magnetoresistive effect with a bump is persistently observed for defective graphene device at various temperatures, while a smooth and symmetric signal is detected for non-defective graphene spin valve. Parallel and antiparallel alignments of magnetization of magnetic materials shows low and high resistance states, respectively. The magnetoresistance (MR) ratio for defective graphene NiFe/graphene/NiFe spin valve is measured to be ~0.16% at 300 K which progresses to ~0.39% for non-defective graphene device at the same temperature. Similarly at 4.2 K the MR ratios are reported to be ~0.41% and ~0.78% for defective and non-defective graphene devices, respectively. Our investigation provides an evidence for relatively better response of the spin valve signal with high quality graphene interlayer.

  1. Heart Valve Surgery Recovery and Follow Up

    Science.gov (United States)

    ... Guide: Understanding Your Heart Valve Problem | Spanish Symptom Tracker | Spanish Pre-surgery Checklist | Spanish What Is Heart ... Heart Valves • Heart Valve Problems and Causes • Risks, Signs and Symptoms • Accurate Diagnosis • Treatment Options • Recovery and ...

  2. ABO blood groups: A risk factor for left atrial and left atrial appendage thrombogenic milieu in patients with non-valvular atrial fibrillation.

    Science.gov (United States)

    Fu, Yuan; Li, Kuibao; Yang, Xinchun

    2017-08-01

    Previous studies have identified ABO blood groups as predictors of thromboembolic diseases. In patients with atrial fibrillation (AF), however, potential association between ABO blood groups and the risk of left atrial (LA) and/or left atrial appendage (LAA) thrombogenic milieu (TM) has not been established. This is a retrospective case-control study that included 125 consecutive patients with non-valvular atrial fibrillation (NVAF) plus TM, as evidenced by transesophageal echocardiography (TEE) during a period from1 January 2010 to 31 December 2016. The controls were selected randomly from 1072 NVAF without TM at a 1:2 ratio. Potential association between ABO blood groups and TM was analyzed using multivariate logistic regression analysis. The risk of TM was higher in patients with blood group A (33.6% vs. 20.2% in non-A blood groups, P=0.005). After adjusting for age, sex, oral anticoagulant use, AF type and duration, and relevant functional measures (e.g., NT-pro BNP level, left atrium diameter, and left ventricular ejection fraction), blood group A remained associated with an increased risk of TM (OR=2.99, 95% CI 1.4-6.388, P=0.005). Blood group A is an independent risk factor for TM in NVAF patients. Copyright © 2017 Elsevier Ltd. All rights reserved.

  3. Simvastatin Attenuates the Oxidative Stress, Endothelial Thrombogenicity and the Inducibility of Atrial Fibrillation in a Rat Model of Ischemic Heart Failure

    Directory of Open Access Journals (Sweden)

    Kyoung-Im Cho

    2014-08-01

    Full Text Available Increased atrial oxidative stress has an important role in inducing and maintaining atrial fibrillation (AF, and the activation of the small GTPase Rac1 contributes to the oxidative stress. We investigated the relationship of Rac1, atrial endothelial thromboprotective markers and AF inducibility and if simvastatin has a potential beneficial effect on a myocardial infarction (MI-induced heart failure (HF rat model. Rats were randomized into three groups (shams, MI group and simvastatin treatment group and underwent echocardiography, AF induction studies and left atrial (LA fibrosis analysis. Atrial Rac 1, sodium calcium exchanger (INCX, sarcoplasmic reticulum calcium ATPase (SERCA, endothelial nitric oxide synthase (eNOS and induced nitric oxide synthase (iNOS were measured. AF inducibility, AF duration and LA fibrosis were significantly higher in the MI group (p < 0.001 vs. sham, which were significantly reduced by simvastatin (p < 0.05 vs. MI. The reduced expressions of atrial eNOS, SERCA, thrombomodulin, tissue factor pathway inhibitor and tissue plasminogen activator in the MI group were significantly improved by simvastatin. Furthermore, the increased expression of atrial iNOS, INCX and Rac1 activity were significantly decreased by the simvastatin. Oxidative stress, endothelial dysfunction and thrombogenicity are associated with the promotion of AF in a rat model of ischemic HF. These were associated with increased Rac1 activity, and simvastatin treatment prevents these changes.

  4. Experimental substantiation of the design of a prosthetic heart valve for «valve-in-valve» implantation

    Directory of Open Access Journals (Sweden)

    K. Yu. Klyshnikov

    2017-01-01

    Full Text Available The aim of the study was to perform a series of in vitro tests of a prototype of the developing heart valve prosthesis to evaluate its functional characteristics. Materials and methods. In this work we have used the frames and full prototypes of the prosthesis, consisting of a stent-like stainless steel support frame with mounted biological leaflets and cover. The authors evaluated the calculated and experimental forces necessary for the displacement of the sutureless implanted prosthesis using the test machine under uniaxial tension. The risk of defects and damages to the supporting framework as a result of implantation was evaluated by scanning electron microscopy. The hydrodynamic characteristics of the prosthesis were investigated under physiological conditions and «valvein-valve» implantation. Evaluation of the ergonomics and applicability of the proposed construction on the cadaver heart model of cattle was carried out. Results. As a result of the forces assessment, it was found that the force required to shear the prosthesis was 3.12 ± 0.37 N, while the calculated value was 1.7 N, which is significantly lower than the obtained value. The comparison of the images obtained with small and large magnifications demonstrated the absence of critical surface defects. Additional analysis under the super-large magnifications also did not reveal problem areas. During the hydrodynamic study, it was shown that the average transplant gradient increased slightly from 2.8–3.4 to 3.2–4.5 mm Hg for the initial prosthesis and the «valve-in-valve» complex, respectively. The decrease of the effective orifice area was 6–9% relative to the initial one. Evaluation of the implantation technique demonstrated the consistency of the approach: the use of the developed holder in combination with the balloon implantation system made it possible to position the prosthesis throughout the procedure. Conclusion. The series of tests demonstrates the consistency

  5. Oil Stiction in Fast Switching Annular Seat Valves for Digital Displacement Fluid Power Machines

    DEFF Research Database (Denmark)

    Roemer, Daniel Beck; Johansen, Per; Pedersen, Henrik C.

    2014-01-01

    Digital Displacement (DD) fluid power machines utilizes electronically controlled seat valves connected to pressure chambers to obtain variable displacement with high operational efficiency and high bandwidth. To achieve high efficiency, fast valve switching is essential and all aspects related...... to the dynamic behaviour of the seat valves must be considered to optimize the machine efficiency. A significant effect influencing the valves switching performance is the presence of oil stiction when separating the contact surfaces in valve opening movement. This oil stiction force is limited by cavitation...... for low pressure levels, e.g. valves connected to the low pressure manifold, however for valves operated at higher pressure levels, the oil stiction force is dominating when the separating surfaces are close to contact. This paper presents an analytic solution to the oil stiction force for annular seat...

  6. A Study on the Main Steam Safety Valve Opening Mechanism by Flashing on NPPs

    International Nuclear Information System (INIS)

    Kim, Bae Joo

    2009-01-01

    A safety injection event happened by opening of the Main Steam Safety Valve at Kori unit 1 on April 16, 2005. The safety valves were opened at the lower system pressure than the valve opening set point due to rapid system pressure drop by opening of the Power Operated Relief Valve installed at the upstream of the Main Steam System. But the opening mechanism of safety valve at the lower set point pressure was not explained exactly. So, it needs to be understood about the safety valve opening mechanism to prevent a recurrence of this kind of event at a similar system of Nuclear Power Plant. This study is aimed to suggest the hydrodynamic mechanism for the safety valve opening at the lower set point pressure and the possibility of the recurrence at similar system conditions through document reviewing for the related previous studies and Kori unit 1 event

  7. ADAM®/SIPLUG®: An innovative valve monitoring system

    International Nuclear Information System (INIS)

    Muñoz, L.; Krell, M.

    2012-01-01

    Optimized maintenance strategies are a key aspect for safe and undisturbed plant operation. Innovative valve service solutions, e.g. valve diagnostics can support this in an efficient way. The ADAM®/SIPLUG® valve monitoring system allows full online monitoring of valves and actuators with automatic evaluation and assessment. Especially for safety-related and operation-related valves this provides valuable information on components condition to ensure proper function and contribute to optimization of maintenance strategies as well as effective maintenance performance. The new SIPLUG®-4 modules are the evolutionary solution for valve diagnosis at the Motor Control Center (MCC). As the SIPLUG®-4 can be installed directly in the MCC outgoing actuator power cable it allows an easy installation in existing switchgear cabinets. Measurement at MCC means also zero effort for performance of diagnostics reducing the number of on-site activities. This results in decrease of maintenance costs and dose rates for deployed personnel. The ADAM® evaluation software and database was developed in parallel with the hardware. It provides automatic analysis of the monitoring results using the limit values specified for the valves. The measured data can be transmitted via the power plant’s local area network to the ADAM® data server, if the SIPLUG® online hardware is installed. With the mobile solution, the data can be transmitted via serial or USB interface to a PC or notebook. With this solution all measurement information will be available immediately in the offices of plant engineers. Also, with SIPLUG® online all operations of valves can be automatically recorded. More than 25 years of experience in various plants worldwide show that the application of ADAM®/SIPLUG® valve diagnostics solution leads to increased plant safety and availability. Some of the references for ADAM®/SIPLUG® are Germany, Switzerland, Brazil, Spain and Eastern Europe. (author)

  8. Pressure locking and thermal binding of gate valves

    Energy Technology Data Exchange (ETDEWEB)

    Kelly, E.M.

    1996-12-01

    Pressure locking and thermal binding represent potential common mode failure mechanisms that can cause safety-related power-operated gate valves to fail in the closed position, thus rendering redundant safety-related systems incapable of performing their safety functions. Supplement 6 to Generic Letter 89-10, {open_quotes}Safety-Related Motor-Operated Gate Valve Testing and Surveillance,{close_quotes} provided an acceptable approach to addressing pressure locking and thermal binding of gate valves. More recently, the NRC has issued Generic Letter 95-07, {open_quotes}Pressure Locking and Thermal Binding of Safety-Related Power-Operated Gate Valves,{close_quotes} to request that licensees take certain actions to ensure that safety-related power-operated gate valves that are susceptible to pressure locking or thermal binding are capable of performing their safety functions within the current licensing bases. Over the past two years, several plants in Region I determined that valves in certain systems were potentially susceptible to pressure locking and thermal binding, and have taken various corrective actions. The NRC Region I Systems Engineering Branch has been actively involved in the inspection of licensee actions in response to the pressure locking and thermal binding issue. Region I continues to maintain an active involvement in this area, including participation with the Office of Nuclear Reactor Regulation in reviewing licensee responses to Generic Letter 95-07.

  9. 241-AN-A valve pit manifold valves and position indication acceptance test procedure

    Energy Technology Data Exchange (ETDEWEB)

    VANDYKE, D.W.

    1999-08-25

    This document describes the method used to test design criteria for gear actuated ball valves installed in 241-AN-A Valve Pit located at 200E Tank Farms. The purpose of this procedure is to demonstrate the following: Equipment is properly installed, labeled, and documented on As-Built drawings; New Manifold Valves in the 241-AN-A Valve Pit are fully operable using the handwheel of the valve operators; New valve position indicators on the valve operators will show correct valve positions; New valve position switches will function properly; and New valve locking devices function properly.

  10. Application of artificial intelligence to motor operated valve testing

    International Nuclear Information System (INIS)

    Bogard, T.; Bednar, F.; Matty, T.; Kent, R.

    1989-01-01

    Improper valve maintenance can be a significant roadblock to successful power plant operation. There have been events during which motor operated valves failed on demand due to improper switch settings. For nuclear electric generating stations, these events have led to regulatory requirements such as NRC Bulletin 85-03 and NRC Bulletin 89-10 Safety Related Motor Operated Valve Testing and Surveillance which imposes strict testing and programmatic requirements on motor operated valves (MOV). Part of the requirements include performing diagnostic testing to verify stem thrust loads and switch settings. Diagnostic equipment must be non-intrusive, minimize valve disassembly, and reduce plant refueling critical path time for testing. In this paper an on-line diagnostic system using sensors to measure stem forces, motor current, and valve position, and a portable system employing these same sensor inputs in addition to torque, limit and torque bypass switch inputs is described. Sophisticated graphic software is employed to display data or trace information. A rule based artificial intelligence (AI) system is used to analyze sensor outputs. Objectives for valve diagnostics, sample AI rules, results of actual field testing, and system software/hardware architecture are presented

  11. A Systematic Review of Infective Endocarditis in Patients With Bovine Jugular Vein Valves Compared With Other Valve Types.

    Science.gov (United States)

    Sharma, Ashutosh; Cote, Anita T; Hosking, Martin C K; Harris, Kevin C

    2017-07-24

    The aim of this study was to systematically evaluate the incidence of infective endocarditis (IE) in right ventricle-to-pulmonary artery conduits and valves, comparing bovine jugular vein (BJV) valves with all others. Recent evidence suggests that the incidence of IE is higher in patients with congenital heart disease who have undergone implantation of BJV valves in the pulmonary position compared with other valves. Systematic searches of published research were conducted using electronic databases (MEDLINE, Embase, and CINAHL) and citations cross-referenced current to April 2016. Included studies met the following criteria: patients had undergone right ventricle-to-pulmonary artery conduit or percutaneous pulmonary valve implantation, and investigators reported on the type of conduit or valve implanted, method of intervention (surgery or catheter based), IE incidence, and follow-up time. Fifty studies (Levels of Evidence: 2 to 4) were identified involving 7,063 patients. The median cumulative incidence of IE was higher for BJV compared with other valves (5.4% vs. 1.2%; p endocarditis with BJV valves than other types of right ventricle-to-pulmonary artery conduits. There was no difference in the incidence of endocarditis between catheter-based bovine valves and surgically implanted bovine valves, suggesting that the substrate for future infection is related to the tissue rather than the method of implantation. Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  12. Aortic valve replacement

    DEFF Research Database (Denmark)

    Kapetanakis, Emmanouil I; Athanasiou, Thanos; Mestres, Carlos A

    2008-01-01

    mortality were collected. Group analysis by patient geographic distribution and by annular diameter of the prosthesis utilized was conducted. Patients with a manufacturer's labeled prosthesis size > or = 21 mm were assigned to the 'large' aortic size subset, while those with a prosthesis size ... differences in the distribution of either gender or BSA. In the multivariable model, south European patients were seven times more likely to receive a smaller-sized aortic valve (OR = 6.5, 95% CI = 4.82-8.83, p

  13. A comparison of conventional surgery, transcatheter aortic valve replacement, and sutureless valves in "real-world" patients with aortic stenosis and intermediate- to high-risk profile.

    Science.gov (United States)

    Muneretto, Claudio; Alfieri, Ottavio; Cesana, Bruno Mario; Bisleri, Gianluigi; De Bonis, Michele; Di Bartolomeo, Roberto; Savini, Carlo; Folesani, Gianluca; Di Bacco, Lorenzo; Rambaldini, Manfredo; Maureira, Juan Pablo; Laborde, Francois; Tespili, Maurizio; Repossini, Alberto; Folliguet, Thierry

    2015-12-01

    We sought to investigate the clinical outcomes of patients with isolated severe aortic stenosis and an intermediate- to high-risk profile treated by means of conventional surgery (surgical aortic valve replacement), sutureless valve implantation, or transcatheter aortic valve replacement in a multicenter evaluation. Among 991 consecutive patients with isolated severe aortic stenosis and an intermediate- to high-risk profile (Society of Thoracic Surgeons score >4 and logistic European System for Cardiac Operative Risk Evaluation I >10), a propensity score analysis was performed on the basis of the therapeutic strategy: surgical aortic valve replacement (n = 204), sutureless valve implantation (n = 204), and transcatheter aortic valve replacement (n = 204). Primary end points were 30-day mortality and overall survival at 24-month follow-up; the secondary end point was survival free from a composite end point of major adverse cardiac events (defined as cardiac-related mortality, myocardial infarction, cerebrovascular accidents, and major hemorrhagic events) and periprosthetic regurgitation greater than 2. Thirty-day mortality was significantly higher in the transcatheter aortic valve replacement group (surgical aortic valve replacement = 3.4% vs sutureless = 5.8% vs transcatheter aortic valve replacement = 9.8%; P = .005). The incidence of postprocedural was 3.9% in asurgical aortic valve replacement vs 9.8% in sutureless vs 14.7% in transcatheter aortic valve replacement (Prisk factor for overall mortality hazard ratio (hazard ratio, 2.5; confidence interval, 1.1-4.2; P = .018). The use of transcatheter aortic valve replacement in patients with an intermediate- to high-risk profile was associated with a significantly higher incidence of perioperative complications and decreased survival at short- and mid-term when compared with conventional surgery and sutureless valve implantation. Copyright © 2015 The American Association for Thoracic Surgery. Published by

  14. BNGS B valve packing program

    International Nuclear Information System (INIS)

    Cumming, D.

    1995-01-01

    The Bruce B Valve Packing Program began in 1987. The early history and development were presented at the 1992 International CANDU Maintenance conference. This presentation covers the evolution of the Bruce B Valve Packing Program over the period 1992 to 1995. (author)

  15. Experience with valves for PHWR reactors

    International Nuclear Information System (INIS)

    Narayan, K.; Mhetre, S.G.

    1977-01-01

    Material specifications and inspection and testing requirements of the valves meant for use in nuclear reactors are mentioned. In the heavy water systems (both primary and moderator) of a PHWR type reactor, the valves used are gate valves, globe valves, diaphragm valves, butterfly valves, check valves and relief valves. Their locations and functions they perform in the Rajasthan Atomic Power Station Unit-1 are described. Experience with them is given. The major problems encountered with them have been : (1) leakage from the stem seals and body bonnet joint, (2) leakage due to failure of diaphragm and/or washout of the packing and (3) malfunctioning. Measures taken to solve these are discussed. Finally a mention has been made of improved versions of valves, namely, metal diaphragm valve and inverted relief valve. (M.G.B.)

  16. Ahmed glaucoma valve implant: surgical technique and complications

    Directory of Open Access Journals (Sweden)

    Riva I

    2017-02-01

    Full Text Available Ivano Riva,1 Gloria Roberti,1 Francesco Oddone,1 Anastasios GP Konstas,2 Luciano Quaranta3 1IRCCS “Fondazione GB Bietti per l’Oftalmologia”, Rome, Italy; 21st University Department of Ophthalmology, Glaucoma Unit, AHEPA Hospital, Thessaloniki, Greece; 3Department of Medical and Surgical Specialties, Section of Ophthalmology, University of Brescia, Brescia, Italy Abstract: Implantation of Ahmed glaucoma valve is an effective surgical technique to reduce intraocular pressure in patients affected with glaucoma. While in the past, the use of this device was reserved to glaucoma refractory to multiple filtration surgical procedures, up-to-date mounting experience has encouraged its use also as a primary surgery for selected cases. Implantation of Ahmed glaucoma valve can be challenging for the surgeon, especially in patients who already underwent previous multiple surgeries. Several tips have to be acquired by the surgeon, and a long learning curve is always needed. Although the valve mechanism embedded in the Ahmed glaucoma valve decreases the risk of postoperative hypotony-related complications, it does not avoid the need of a careful follow-up. Complications related to this type of surgery include early and late postoperative hypotony, excessive capsule fibrosis around the plate, erosion of the tube or plate edge, and very rarely infection. The aim of this review is to describe surgical technique for Ahmed glaucoma valve implantation and to report related complications. Keywords: glaucoma, surgical technique, glaucoma drainage devices, Ahmed glaucoma valve, complications

  17. Prosthetic valve endocarditis after transcatheter aortic valve implantation

    DEFF Research Database (Denmark)

    Olsen, Niels Thue; De Backer, Ole; Thyregod, Hans G H

    2015-01-01

    BACKGROUND: Transcatheter aortic valve implantation (TAVI) is an advancing mode of treatment for inoperable or high-risk patients with aortic stenosis. Prosthetic valve endocarditis (PVE) after TAVI is a serious complication, but only limited data exist on its incidence, outcome, and procedural......%) were treated conservatively and 1 with surgery. Four patients (22%) died from endocarditis or complications to treatment, 2 of those (11%) during initial hospitalization for PVE. An increased risk of TAVI-PVE was seen in patients with low implanted valve position (hazard ratio, 2.8 [1.1-7.2]), moderate...

  18. A knowledge-based assistant for valve maintenance planning

    International Nuclear Information System (INIS)

    Winter, M.J.; Danofsky, R.A.; Spinrad, B.I.; Howard, K.

    1987-01-01

    A knowledge-based program is being developed to assist engineers in maintenance planning for safety related, motor-operated valves at a boiling water reactor. The purpose of this project is to develop the general framework for a prototype system that demonstrates the capabilities for diagnosing valve symptoms and prescribing corrective maintenance, completing a portion of the Corrective Maintenance Action Request (CMAR) form which must be prepared for each job, and managing an interactive valve data base. Minimizing user input and providing output in a form that is familiar to the maintenance planning engineer are important goals for the program. This paper describes the present features of the valve maintenance advisory system which is currently being tested

  19. Cavitation guide for control valves

    Energy Technology Data Exchange (ETDEWEB)

    Tullis, J.P. [Tullis Engineering Consultants, Logan, UT (United States)

    1993-04-01

    This guide teaches the basic fundamentals of cavitation to provide the reader with an understanding of what causes cavitation, when it occurs, and the potential problems cavitation can cause to a valve and piping system. The document provides guidelines for understanding how to reduce the cavitation and/or select control valves for a cavitating system. The guide provides a method for predicting the cavitation intensity of control valves, and how the effect of cavitation on a system will vary with valve type, valve function, valve size, operating pressure, duration of operation and details of the piping installation. The guide defines six cavitation limits identifying cavitation intensities ranging from inception to the maximum intensity possible. The intensity of the cavitation at each limit Is described, including a brief discussion of how each level of cavitation influences the valve and system. Examples are included to demonstrate how to apply the method, including making both size and pressure scale effects corrections. Methods of controlling cavitation are discussed providing information on various techniques which can be used to design a new system or modify an existing one so it can operate at a desired level of cavitation.

  20. Cavitation guide for control valves

    International Nuclear Information System (INIS)

    Tullis, J.P.

    1993-04-01

    This guide teaches the basic fundamentals of cavitation to provide the reader with an understanding of what causes cavitation, when it occurs, and the potential problems cavitation can cause to a valve and piping system. The document provides guidelines for understanding how to reduce the cavitation and/or select control valves for a cavitating system. The guide provides a method for predicting the cavitation intensity of control valves, and how the effect of cavitation on a system will vary with valve type, valve function, valve size, operating pressure, duration of operation and details of the piping installation. The guide defines six cavitation limits identifying cavitation intensities ranging from inception to the maximum intensity possible. The intensity of the cavitation at each limit Is described, including a brief discussion of how each level of cavitation influences the valve and system. Examples are included to demonstrate how to apply the method, including making both size and pressure scale effects corrections. Methods of controlling cavitation are discussed providing information on various techniques which can be used to design a new system or modify an existing one so it can operate at a desired level of cavitation

  1. Characteristic analysis of servo valve

    International Nuclear Information System (INIS)

    Ko, J. H.; Ryu, D. R.; Lee, J. H.; Kim, Y. S.; Na, J. C.; Kim, D. S.

    2008-01-01

    Electro-pneumatic servo valve is an electro-mechanical device which converts electric signals into a proper pneumatic flow rate or pressure. In order to improve the overall performance of pneumatic servo systems, electro-pneumatic servo valves are required, which have fast dynamic characteristics, no air leakage at a null point, and can be fabricated at a low-cost. The first objective of this research is to design and to fabricate a new electro-pneumatic servo valve which satisfies the above-mentioned requirements. In order to design the mechanism of the servo valve optimally, the flow inside the valve depending upon the position of spool was analyzed variously, and on the basis of such analysis results, the valve mechanism, which was formed by combination of the spool and the sleeve, was designed and manufactured. And a tester for conducting an overall performance test was designed and manufactured, and as a result of conducting the flow rate test, the pressure test and the frequency test on the developed pneumatic servo valve

  2. Proving test on the reliability for nuclear valves

    International Nuclear Information System (INIS)

    Kajiyama, Yasuo; Tashiro, Hisao; Uga, Takeo; Maeda, Shunichi.

    1986-01-01

    Since valves are the most common components, they could be the most frequent causes of troubles in nuclear power plants. This proving test, therefore, has an important meaning to examine and verify the reliability of various valves under simulating conditions of abnormal and transient operations of the nuclear power plant. The test was performed mainly for the various types and pressure ratings of valves which were used in the primary and secondary systems in BWR and PWR nuclear power plants and which had major operating or safety related functions in those nuclear power plants. The results of the proving test, confirmed for more than four years, showed relatively favourable performance of the tested valves. It is concluded that performances of valves including operability, seat sealing and structural integrity were proved under the thermal cycling, vibration and pipe reaction load conditions. Operating functions during and after accident such as loss of coolant accident were satisfactory. From these results, it was considered that the purpose of this proving test was satisfactorily fulfilled. Several data accumulated by the test would be useful to get better reliability if it was evaluated with the actually experienced data of valves in the nuclear power plants. (Nogami, K.)

  3. System for remotely servicing a top loading captive ball valve

    International Nuclear Information System (INIS)

    Berry, S.M.; Porter, M.L.

    1996-01-01

    An attachment for facilitating servicing of a valve is disclosed including: an assembly composed of a valve seat defining a flow path, a flow control member movable relative to the valve seat for blocking or unblocking the valve seat, and a control device including a stem coupled to the flow control member and operable for moving the flow control member relative to the valve se housing for receiving the assembly, the housing having an opening via which the assembly can be removed from, and installed in, the housing, and the housing having a plurality of threaded studs which surround the opening and project away from the housing; a valve housing cover for closing and sealing the opening in the housing, the cover having a first bore for passage of the stem of the control device when the assembly is installed in the housing and a plurality of second bores each located for passage of a respective stud when the cover closes the opening in the housing. A plurality of threaded nuts are engageable with the studs for securing the cover to the housing when the cover closes the opening in the housing, wherein the attachment comprises: a plurality of nut guide devices removable from the housing and each operatively associated with a respective stud for retaining a respective nut and guiding the respective nut into alignment with the respective stud to enable the respective nut to be rotated into engagement with the respective stud; and aligning the nut guide devices with the studs. 7 figs

  4. Developments in mechanical heart valve prosthesis

    Indian Academy of Sciences (India)

    Artificial heart valves are engineered devices used for replacing diseased or damaged natural valves of the heart. Most commonly used for replacement are mechanical heart valves and biological valves. This paper briefly outlines the evolution, designs employed, materials being used,. and important factors that affect the ...

  5. Numerical Analysis of Combined Valve Hydrodynamic Characteristics for Turbine System

    Energy Technology Data Exchange (ETDEWEB)

    Bhowmik, P. K.; Shamim, J. A.; Gairola, A.; Arif, M.; Suh, Kune Y. [Seoul National Univ., Seoul (Korea, Republic of)

    2014-05-15

    precisely by the valve manufacturer. As a matter of fact, attempts were made to obtain flow characteristic curves resorting to analytical as well as numerical methods. The flow characteristic curve relates the stem lift with mass flow rate at a specific temperature and pressure. This paper focuses on computational and numerical analysis of the combined stop and control valve. Combined Airflow Regulation Analysis (CARA) is performed to check on the hydrodynamic characteristic, which is represented by flow coefficient characteristic. CATIA V.5 and ANSYS CFX are used for three-dimensional computer-aided design and computational fluid dynamics (CFD) analysis, respectively. Flow characteristic curves are plotted by calculating theoretical and numerical mass flow rate. Hydrodynamic analysis was made of the combined stop and control valve for the turbine system using ANSYS CFX. The result of the numerical study represented by the valve flow coefficient with different normalized values of valve stem movement L/D and different pressure ratios of valve outlet and inlet agrees well with the ideal case and other similar previous experimental results. This study also provided a solid understanding with versatile options for analyzing the hydrodynamics of the combined valve considering the various internal geometry, seat, plug, and the inlet plus outlet boundary conditions to improve the efficiency, performance and reliability of the turbine system of small as well as large power conversion system using the numerical analysis with minimal cost and time.

  6. Numerical Analysis of Combined Valve Hydrodynamic Characteristics for Turbine System

    International Nuclear Information System (INIS)

    Bhowmik, P. K.; Shamim, J. A.; Gairola, A.; Arif, M.; Suh, Kune Y.

    2014-01-01

    precisely by the valve manufacturer. As a matter of fact, attempts were made to obtain flow characteristic curves resorting to analytical as well as numerical methods. The flow characteristic curve relates the stem lift with mass flow rate at a specific temperature and pressure. This paper focuses on computational and numerical analysis of the combined stop and control valve. Combined Airflow Regulation Analysis (CARA) is performed to check on the hydrodynamic characteristic, which is represented by flow coefficient characteristic. CATIA V.5 and ANSYS CFX are used for three-dimensional computer-aided design and computational fluid dynamics (CFD) analysis, respectively. Flow characteristic curves are plotted by calculating theoretical and numerical mass flow rate. Hydrodynamic analysis was made of the combined stop and control valve for the turbine system using ANSYS CFX. The result of the numerical study represented by the valve flow coefficient with different normalized values of valve stem movement L/D and different pressure ratios of valve outlet and inlet agrees well with the ideal case and other similar previous experimental results. This study also provided a solid understanding with versatile options for analyzing the hydrodynamics of the combined valve considering the various internal geometry, seat, plug, and the inlet plus outlet boundary conditions to improve the efficiency, performance and reliability of the turbine system of small as well as large power conversion system using the numerical analysis with minimal cost and time

  7. Butterfly valve torque prediction methodology

    International Nuclear Information System (INIS)

    Eldiwany, B.H.; Sharma, V.; Kalsi, M.S.; Wolfe, K.

    1994-01-01

    As part of the Motor-Operated Valve (MOV) Performance Prediction Program, the Electric Power Research Institute has sponsored the development of methodologies for predicting thrust and torque requirements of gate, globe, and butterfly MOVs. This paper presents the methodology that will be used by utilities to calculate the dynamic torque requirements for butterfly valves. The total dynamic torque at any disc position is the sum of the hydrodynamic torque, bearing torque (which is induced by the hydrodynamic force), as well as other small torque components (such as packing torque). The hydrodynamic torque on the valve disc, caused by the fluid flow through the valve, depends on the disc angle, flow velocity, upstream flow disturbances, disc shape, and the disc aspect ratio. The butterfly valve model provides sets of nondimensional flow and torque coefficients that can be used to predict flow rate and hydrodynamic torque throughout the disc stroke and to calculate the required actuation torque and the maximum transmitted torque throughout the opening and closing stroke. The scope of the model includes symmetric and nonsymmetric discs of different shapes and aspects ratios in compressible and incompressible fluid applications under both choked and nonchoked flow conditions. The model features were validated against test data from a comprehensive flowloop and in situ test program. These tests were designed to systematically address the effect of the following parameters on the required torque: valve size, disc shapes and disc aspect ratios, upstream elbow orientation and its proximity, and flow conditions. The applicability of the nondimensional coefficients to valves of different sizes was validated by performing tests on 42-in. valve and a precisely scaled 6-in. model. The butterfly valve model torque predictions were found to bound test data from the flow-loop and in situ testing, as shown in the examples provided in this paper

  8. Deep venous thrombosis: The valve cusp hypoxia thesis and its incompatibility with modern orthodoxy.

    Science.gov (United States)

    Malone, P Colm; Agutter, Paul S

    2016-01-01

    The valve cusp hypoxia thesis (VCHT) of the aetiology of deep venous thrombosis (DVT) was adumbrated in this journal in 1977 and fully articulated in 2008, the original hypothesis having been strongly corroborated by experiments published in 1981 and 1984. It presents a unitary account of the pathogenesis of venous thrombosis and embolism that is rooted in the pathophysiological tradition of Hunter, Virchow, Lister, Welch and Aschoff, a tradition traceable back to Harvey. In this paper we summarise the thesis in its mature form, consider its compatibility with recent advances in the DVT field, and ask why it has not yet been assimilated into the mainstream literature, which during the past half century has been dominated by a haematology-orientated 'consensus model'. We identify and discuss seven ways in which the VCHT is incompatible with these mainstream beliefs about the aetiology of venous thrombosis, drawing attention to: (1) the spurious nature of 'Virchow's triad'; (2) the crucial differences between 'venous thrombus' and 'clot'; the facts that (3) venous thrombi form in the valve pockets (VVPs), (4) DVT is not a primarily haematological condition, (5) the so-called 'thrombophilias' are not thrombogenic per se; (6) the conflict between the single unitary aetiology of DVT and the tacit assumption that the condition is 'multicausal'; (7) the inability of anticoagulants to prevent the initiation of venous thrombogenesis, though they do prevent the growth of thrombi to clinically significant size. In discussing point (7), we show that the VCHT indicates new approaches to mechanical prophylaxis against DVT. These approaches are then formulated as experimentally testable hypotheses, and we suggest methods for testing them preclinically using animal trials. Copyright © 2015 Elsevier Ltd. All rights reserved.

  9. Early spontaneous intermittent myocardial reperfusion during acute myocardial infarction is associated with augmented thrombogenic activity and less myocardial damage

    NARCIS (Netherlands)

    Haider, A.W.; Andreotti, F.; Hackett, D.R.; Tousoulis, D.; Kluft, C.; Maseri, A.; Davies, G.J.

    1995-01-01

    Objectives. This study investigated the influence of early spontaneous intermittent reperfusion on the extent of myocardial damage and its relation to endogenous hemostatic activity, Background. In the early phase of acute myocardial infarction coronary occlusion is often intermittent, even before

  10. Thermostatic Radiator Valve Evaluation

    Energy Technology Data Exchange (ETDEWEB)

    Dentz, J. [Advanced Residential Integrated Energy Solutions Collaborative (ARIES), New York, NY (United States); Ansanelli, E. [Advanced Residential Integrated Energy Solutions Collaborative (ARIES), New York, NY (United States)

    2015-01-01

    A large stock of multifamily buildings in the Northeast and Midwest are heated by steam distribution systems. Losses from these systems are typically high and a significant number of apartments are overheated much of the time. Thermostatically controlled radiator valves (TRVs) are one potential strategy to combat this problem, but have not been widely accepted by the residential retrofit market. In this project, the ARIES team sought to better understand the current usage of TRVs by key market players in steam and hot water heating and to conduct limited experiments on the effectiveness of new and old TRVs as a means of controlling space temperatures and reducing heating fuel consumption. The project included a survey of industry professionals, a field experiment comparing old and new TRVs, and cost-benefit modeling analysis using BEopt™ (Building Energy Optimization software).

  11. The Diagnosis of Internal Leakage of Control Valve Based on the Grey Correlation Analysis Method

    Directory of Open Access Journals (Sweden)

    Zheng DING

    2014-07-01

    Full Text Available The valve plays an important part in the industrial automation system. Whether it operates normally or not relates with the quality of the products directly while its faults are relatively common because of bad working conditions. And the internal leakage is one of the common faults. Consequently, this paper sets up the experimental platform to make the valve in different working condition and collect relevant data online. Then, diagnose the internal leakage of the valve by using the grey correlation analysis method. The results show that this method can not only diagnose the internal leakage of valve accurately, but also distinguish fault degree quantitatively.

  12. Emerging Trends in Heart Valve Engineering: Part IV. Computational Modeling and Experimental Studies.

    Science.gov (United States)

    Kheradvar, Arash; Groves, Elliott M; Falahatpisheh, Ahmad; Mofrad, Mohammad K; Hamed Alavi, S; Tranquillo, Robert; Dasi, Lakshmi P; Simmons, Craig A; Jane Grande-Allen, K; Goergen, Craig J; Baaijens, Frank; Little, Stephen H; Canic, Suncica; Griffith, Boyce

    2015-10-01

    In this final portion of an extensive review of heart valve engineering, we focus on the computational methods and experimental studies related to heart valves. The discussion begins with a thorough review of computational modeling and the governing equations of fluid and structural interaction. We then move onto multiscale and disease specific modeling. Finally, advanced methods related to in vitro testing of the heart valves are reviewed. This section of the review series is intended to illustrate application of computational methods and experimental studies and their interrelation for studying heart valves.

  13. Valve leakage inspection testing and maintenance process

    International Nuclear Information System (INIS)

    Aikin, J.A.; Reinwald, J.W.; Kittmer, C.A.

    1991-01-01

    In valve maintenance, packing rings that prevent leakage along the valve stem must periodically be replaced, either during routine maintenance or to correct a leak or valve malfunction. Tools and procedures currently in use for valve packing removal and inspection are generally of limited value due to various access and application problems. A process has been developed by AECL Research that addresses these problems. The process, using incompressible fluid pressure, quickly and efficiently confirms the integrity of the valve backseat, extracts hard-to-remove valve packing sets, and verifies the leak tightness of the repacked valve

  14. Transcatheter, valve-in-valve transapical aortic and mitral valve implantation, in a high risk patient with aortic and mitral prosthetic valve stenoses

    Directory of Open Access Journals (Sweden)

    Harish Ramakrishna

    2015-01-01

    Full Text Available Transcatheter valve implantation continues to grow worldwide and has been used principally for the nonsurgical management of native aortic valvular disease-as a potentially less invasive method of valve replacement in high-risk and inoperable patients with severe aortic valve stenosis. Given the burden of valvular heart disease in the general population and the increasing numbers of patients who have had previous valve operations, we are now seeing a growing number of high-risk patients presenting with prosthetic valve stenosis, who are not potential surgical candidates. For this high-risk subset transcatheter valve delivery may be the only option. Here, we present an inoperable patient with severe, prosthetic valve aortic and mitral stenosis who was successfully treated with a trans catheter based approach, with a valve-in-valve implantation procedure of both aortic and mitral valves.

  15. Tomographic PIV behind a prosthetic heart valve

    Science.gov (United States)

    Hasler, D.; Landolt, A.; Obrist, D.

    2016-05-01

    The instantaneous three-dimensional velocity field past a bioprosthetic heart valve was measured using tomographic particle image velocimetry. Two digital cameras were used together with a mirror setup to record PIV images from four different angles. Measurements were conducted in a transparent silicone phantom with a simplified geometry of the aortic root. The refraction indices of the silicone phantom and the working fluid were matched to minimize optical distortion from the flow field to the cameras. The silicone phantom of the aorta was integrated in a flow loop driven by a piston pump. Measurements were conducted for steady and pulsatile flow conditions. Results of the instantaneous, ensemble and phase-averaged flow field are presented. The three-dimensional velocity field reveals a flow topology, which can be related to features of the aortic valve prosthesis.

  16. Valve-aided twisted Savonius rotor

    Energy Technology Data Exchange (ETDEWEB)

    Jaya Rajkumar, M.; Saha, U.K.

    2006-05-15

    Accessories, such as end plates, deflecting plates, shielding and guide vanes, may increase the power of a Savonius rotor, but make the system structurally complex. In such cases, the rotor can develop a relatively large torque at small rotational speeds and is cheap to build, however it harnesses only a small fraction of the incident wind energy. Another proposition for increasing specific output is to place non-return valves inside the concave side of the blades. Such methods have been studied experimentally with a twisted-blade Thus improving a Savonius rotor's energy capture. This new concept has been named as the 'Valve-Aided Twisted Savonius'rotor. Tests were conducted in a low-speed wind tunnel to evaluate performance. This mechanism is found to be independent of flow direction, and shows potential for large machines. [Author].

  17. An application of the valve-leak monitoring system to the valves for the improved Korean standard nuclear power plant (KSNP+)

    International Nuclear Information System (INIS)

    Byeong-yeol AHN; Dae-sik CHOI; Sang-kook CHUNG

    2006-01-01

    The loss of steam due to valve leakage leads to the inefficiency of power generation at the nuclear power plants. Under the normal conditions of plant operation, it is difficult to detect valve leaks early enough to prevent consequential damages and losses. The capability of timely detection allows the plant adequate time to prepare repair plans, which would ultimately result in efficient power production. Therefore, timing of detection is one of the most important factors in dealing with valve leakage problems. The VLMS has been developed to meet such an industrial demand. It provides early detection of valve leakage by real-time monitoring through the acoustic sensors installed on the inlet and the outlet of the valve. The KSNP+ utilizes the VLMS to enhance the performance and maintenance of major valves at plants. The VLMS will enable the plant to detect the leakage of valve at an early stage. It can reduce the steam losses and save related valve maintenance cost by performing fast diagnosis of valve leakage. (authors)

  18. Developing an optimal valve closing rule curve for real-time pressure control in pipes

    Energy Technology Data Exchange (ETDEWEB)

    Bazarganlari, Mohammad Reza; Afshar, Hossein [Islamic Azad University, Tehran (Iran, Islamic Republic of); Kerachian, Reza [University of Tehran, Tehran (Iran, Islamic Republic of); Bashiazghadi, Seyyed Nasser [Iran University of Science and Technology, Tehran (Iran, Islamic Republic of)

    2013-01-15

    Sudden valve closure in pipeline systems can cause high pressures that may lead to serious damages. Using an optimal valve closing rule can play an important role in managing extreme pressures in sudden valve closure. In this paper, an optimal closing rule curve is developed using a multi-objective optimization model and Bayesian networks (BNs) for controlling water pressure in valve closure instead of traditional step functions or single linear functions. The method of characteristics is used to simulate transient flow caused by valve closure. Non-dominated sorting genetic algorithms-II is also used to develop a Pareto front among three objectives related to maximum and minimum water pressures, and the amount of water passes through the valve during the valve-closing process. Simulation and optimization processes are usually time-consuming, thus results of the optimization model are used for training the BN. The trained BN is capable of determining optimal real-time closing rules without running costly simulation and optimization models. To demonstrate its efficiency, the proposed methodology is applied to a reservoir-pipe-valve system and the optimal closing rule curve is calculated for the valve. The results of the linear and BN-based valve closure rules show that the latter can significantly reduce the range of variations in water hammer pressures.

  19. The Double-Orifice Valve Technique to Treat Tricuspid Valve Incompetence.

    Science.gov (United States)

    Hetzer, Roland; Javier, Mariano; Delmo Walter, Eva Maria

    2016-01-01

    A straightforward tricuspid valve (TV) repair technique was used to treat either moderate or severe functional (normal valve with dilated annulus) or for primary/organic (Ebstein's anomaly, leaflet retraction/tethering and chordal malposition/tethering, with annular dilatation) TV incompetence, and its long-term outcome assessed. A double-orifice valve technique was employed in 91 patients (mean age 52.6 ± 23.2 years; median age 56 years; range: 0.6-82 years) with severe tricuspid regurgitation. Among the patients, three had post-transplant iatrogenic chordal rupture, five had infective endocarditis, 11 had mitral valve insufficiency, 23 had Ebstein's anomaly, and 47 had isolated severe TV incompetence. The basic principle was to reduce the distance between the coapting leaflets, wherein the most mobile leaflet could coapt to the opposite leaflet, by creating two orifices, ensuring valve competence. The TV repair was performed through a median sternotomy or right anterior thoracotomy in the fifth intercostal space under cardiopulmonary bypass. The degree and extent of creating a double-valve orifice was determined by considering the minimal body surface area (BSA)-related acceptable TV diameter. Repair was accomplished by passing pledgeted mattress sutures from the middle of the true anterior annulus to a spot on the opposite septal annulus, located approximately two-thirds of the length of the septal annulus to avoid injury to the bundle of His. The annular apposition divides the TV into a larger anterior and a smaller posterior orifices, enabling valve closure, on both sides. In adults, the diameter of the anterior valve orifice should be 23-25 mm, and the posterior orifice 15-18 mm; thus, the total valve orifice area is 5-6 cm2. In children, the total valve orifice should be a standard deviation of 1.7 mm for a BSA of 1.0m2. During a mean follow up of 8.7 ± 1.34 years (median 10 years; range: 1.5-25.9 years) there have been no reoperations for TV insufficiency

  20. Options for Heart Valve Replacement

    Science.gov (United States)

    ... Guide: Understanding Your Heart Valve Problem | Spanish Symptom Tracker | Spanish Pre-surgery Checklist | Spanish What Is Heart ... Cardiac Arrest: How Are They Different? 7 Warning Signs of a Heart Attack 8 Low Blood Pressure - ...

  1. Minimally invasive aortic valve replacement

    DEFF Research Database (Denmark)

    Foghsgaard, Signe; Schmidt, Thomas Andersen; Kjaergard, Henrik K

    2009-01-01

    In this descriptive prospective study, we evaluate the outcomes of surgery in 98 patients who were scheduled to undergo minimally invasive aortic valve replacement. These patients were compared with a group of 50 patients who underwent scheduled aortic valve replacement through a full sternotomy...... operations were completed as mini-sternotomies, 4 died later of noncardiac causes. The aortic cross-clamp and perfusion times were significantly different across all groups (P replacement...... is an excellent operation in selected patients, but its true advantages over conventional aortic valve replacement (other than a smaller scar) await evaluation by means of randomized clinical trial. The "extended mini-aortic valve replacement" operation, on the other hand, is a risky procedure that should...

  2. Plant experience with check valves in passive systems

    Energy Technology Data Exchange (ETDEWEB)

    Pahladsingh, R R [GKN Joint Nuclear Power Plant, Dodewaard (Netherlands)

    1996-12-01

    In the design of the advanced nuclear reactors there is a tendency to introduce more passive safety systems. The 25 year old design of the GKN nuclear reactor is different from the present BWR reactors because of some special features, such as the Natural Circulation - and the Passive Isolation Condenser system. When reviewing the design, one can conclude that the plant has 25 years of experience with check valves in passive systems and as passive components in systems. The result of this experience has been modeled in a plant-specific ``living PSA`` for the plant. A data-analysis has been performed on components which are related to the safety systems in the plant. As part of this study also the check valves have been taken in consideration. At GKN, the check valves have shown to be reliable components in the systems and no catastrophic failures have been experienced during the 25 years of operation. Especially the Isolation Condenser with its operation experience can contribute substantially to the insight of check valves in stand-by position at reactor pressure and operating by gravity under different pressure conditions. With the introduction of several passive systems in the SBWR-600 design, such as the Isolation Condensers, Gravity Driven Cooling, and Suppression Pool Cooling System, the issue of reliability of check valves in these systems is actual. Some critical aspects for study in connection with check valves are: What is the reliability of a check valve in a system at reactor pressure, to open on demand; what is the reliability of a check valve in a system at low pressure (gravity), to open on demand; what is the reliability of a check valve to open/close when the stand-by check wave is at zero differential pressure. The plant experience with check valves in a few essential safety systems is described and a brief introduction will be made about the application of check valves in the design of the new generation reactors is given. (author). 6 figs, 1 tab.

  3. A nuclear radiation actuated valve for a nuclear reactor

    International Nuclear Information System (INIS)

    Christiansen, D.W.; Schively, D.P.

    1983-01-01

    The valve has a first part (such as a valve rod with piston) and a second part (such as a valve tube surrounding the valve rod, with the valve tube having side slots surrounding the piston). Both valve parts have known nuclear radiation swelling characteristics which are different. The valve parts are positioned so that the valve's first and second parts create a valve orifice which changes in size due to the different nuclear radiation caused swelling of the valve's first part compared to the valve's second part. The valve may be used in a nuclear reactor's core coolant system. (author)

  4. MOVES: A knowledge-based system for maintenance planning for motor-operated valves

    International Nuclear Information System (INIS)

    Winter, M.

    1987-01-01

    Over the past several years, knowledge-based expert systems have emerged as an important part of the general research area known as artificial intelligence. This paper describes a cooperative effort between faculty members at Iowa State University and engineers at the Duane Arnold Energy Center [a 545-MW(electric) boiling water reactor operated by Iowa Electric Light and Power Company] to explore the development of an advisory system for valve maintenance planning. This knowledge-based program, known as Motor-Operated Valves Expert System (MOVES), has a data base that currently includes safety-related motor-operated valves (∼117 valves). Valve maintenance was selected as the subject for the expert system because of the importance of valves in nuclear plant and their impact of plant availability. MOVES is being developed using the microcomputer-(IBM compatible) based expert system tool INSIGHT2+. The authors have found that the project benefits both the university and the utility

  5. Valve Repair Is Superior to Replacement in Most Patients With Coexisting Degenerative Mitral Valve and Coronary Artery Diseases.

    Science.gov (United States)

    Javadikasgari, Hoda; Gillinov, A Marc; Idrees, Jay J; Mihaljevic, Tomislav; Suri, Rakesh M; Raza, Sajjad; Houghtaling, Penny L; Svensson, Lars G; Navia, José L; Mick, Stephanie L; Desai, Milind Y; Sabik, Joseph F; Blackstone, Eugene H

    2017-06-01

    For mitral regurgitation (MR) from degenerative mitral disease in patients with coexisting coronary artery disease, the appropriate surgical strategy remains controversial. From 1985 to 2011, 1,071 adults (age 70 ± 9.3 years, 77% men) underwent combined coronary artery bypass grafting and either mitral valve repair (n = 872, 81%) or replacement (n=199, 19%) for degenerative MR. Propensity matching (177 patient pairs, 89% of possible matches) was used to compare early outcomes and time-related recurrence of MR after mitral valve repair, mitral valve reoperation, and mortality. Risk factors for death were identified with multivariable, multiphase hazard-function analysis. Patients undergoing valve replacement were older, with more valve calcification and a higher prevalence of preoperative atrial fibrillation and heart failure (all p < .0001). Among matched pairs, mitral replacement versus repair was associated with higher hospital mortality (5.0% vs 1.0%, p = .0001) and more postoperative renal failure (7.0% vs 3.2%, p = .01), reexplorations for bleeding (6.0% vs 3.1%, p = .05), and respiratory failure (14% vs 4.7%, p < .0001). Of matched patients undergoing repair, 18% had MR above 3+ by 5 years. Mitral valve durability was similar between matched groups, but survival at 15 years was 18% after replacement versus 52% after repair. Nomograms from the multivariable equation revealed that in 94% of cases, 10-year survival was calculated to be higher after repair than after replacement. In patients with coexisting degenerative mitral valve and coronary artery diseases, mitral valve repair is expected to confer a long-term survival advantage over replacement despite some recurrence of MR. When feasible, it is the procedure of choice for these patients. Copyright © 2017 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  6. Trial design: Rivaroxaban for the prevention of major cardiovascular events after transcatheter aortic valve replacement: Rationale and design of the GALILEO study

    NARCIS (Netherlands)

    Windecker, Stephan; Tijssen, Jan; Giustino, Gennaro; Guimarães, Ana H. C.; Mehran, Roxana; Valgimigli, Marco; Vranckx, Pascal; Welsh, Robert C.; Baber, Usman; van Es, Gerrit-Anne; Wildgoose, Peter; Volkl, Albert A.; Zazula, Ana; Thomitzek, Karen; Hemmrich, Melanie; Dangas, George D.

    2017-01-01

    Optimal antithrombotic treatment after transcatheter aortic valve replacement (TAVR) is unknown and determined empirically. The direct factor Xa inhibitor rivaroxaban may potentially reduce TAVR-related thrombotic complications and premature valve failure. GALILEO is an international, randomized,

  7. Valve performance concept move from preventive to condition-oriented maintenance

    Energy Technology Data Exchange (ETDEWEB)

    Zanner, G.; Kradepohl, P.

    1996-12-01

    As a turnkey supplier of nuclear and fossil power plants, Siemens must pay attention in concentrating, maintaining, and developing the expertise in many areas such as system design, components, materials, quality assurance, and qualification testing within centralized organizations. In the company segment VALVES, Siemens/KWU is staffed with experienced professionals who have serviced the power plant industry for about 25 years. The valve engineers deal with all kinds of valve and actuator-related activities like design ratings, development, qualifications, and ongoing improvements. In this regard, the engineers are involved in nearly all actual problems and suggested solutions through continuing dialogues with utilities, authorities, and vendors of valves and actuators.

  8. AVARIS. An innovative procedure to repair seal seats surfaces in gate and check valves

    International Nuclear Information System (INIS)

    Herzing, Karl-Heinz; Breitenberger, Ulf; Grieser, Armin

    2011-01-01

    AVARIS (AREVA Valve Repair In-Situ) saves the plant operators time and money, as complete exchanges of valves and all related elaborate consequences become dispensable. AVARIS is of high interest in the worldwide maintenance market, as it can be applied in both nuclear and conventional power plants. Moreover, AVARIS retrofits valves in the secondary and primary circuit which even saves complex nuclear transports. AVARIS fulfils the quality requirements of third party authorities in national and international markets. Valves repaired with AVARIS correspond to their original condition concerning sealing function. (orig.)

  9. Valve performance concept move from preventive to condition-oriented maintenance

    International Nuclear Information System (INIS)

    Zanner, G.; Kradepohl, P.

    1996-01-01

    As a turnkey supplier of nuclear and fossil power plants, Siemens must pay attention in concentrating, maintaining, and developing the expertise in many areas such as system design, components, materials, quality assurance, and qualification testing within centralized organizations. In the company segment VALVES, Siemens/KWU is staffed with experienced professionals who have serviced the power plant industry for about 25 years. The valve engineers deal with all kinds of valve and actuator-related activities like design ratings, development, qualifications, and ongoing improvements. In this regard, the engineers are involved in nearly all actual problems and suggested solutions through continuing dialogues with utilities, authorities, and vendors of valves and actuators

  10. Small sodium valve design and operating experience

    International Nuclear Information System (INIS)

    Abramson, R.; Elie, X.; Vercasson, M.; Nedelec, J.

    1974-01-01

    Conventionally, valves for sodium pipes smaller than 125 mm in diameter are called ''small sodium valves''. However, this limit should rather be considered as the lower limit o ''large sodium valves''. In fact, both the largest sizes of small valves and the smallest of large valves can be found in the range of 125-300 mm in diameter. Thus what is said about small valves also applies, for a few valve types, above the 125 mm limit. Sodium valves are described here in a general manner, with no manufacturing details except when necessary for understanding valve behavior. Operating experience is pointed out wherever possible. Finally, some information is given about ongoing or proposed development plans. (U.S.)

  11. Design of the Modular Pneumatic Valve Terminal

    Directory of Open Access Journals (Sweden)

    Jakub E. TAKOSOGLU

    2015-11-01

    Full Text Available The paper presents design of the modular pneumatic valve terminal, which was made on the basis of the patent application No A1 402905 „A valve for controlling fluid power drives, specially for pneumatic actuators, and the control system for fluid power drives valves”. The authors describe a method of operation of the system with double-acting valve and 5/2 (five ways and two position valve. Functions of the valve, and an example of application of the valve terminal in the production process were presented. 3D solid models of all the components of the valve were made. The paper presents a complete 3D model of the valve in various configurations. Using CAD-embedded SOLIDWORKS Flow Simulation computational fluid dynamics CFD analysis was also carried out of compressed air flow in the ways of the valve elements

  12. Development of an effective valve packing program

    Energy Technology Data Exchange (ETDEWEB)

    Hart, K.A.

    1996-12-01

    Current data now shows that graphite valve packing installed within the guidance of a controlled program produces not only reliable stem sealing but predictable running loads. By utilizing recent technological developments in valve performance monitoring for both MOV`s and AOV`s, valve packing performance can be enhanced while reducing maintenance costs. Once known, values are established for acceptable valve packing loads, the measurement of actual valve running loads via the current MOV/AOV diagnostic techniques can provide indication of future valve stem sealing problems, improper valve packing installation or identify the opportunity for valve packing program improvements. At times the full benefit of these advances in material and predictive technology remain under utilized due to simple past misconceptions associated with valve packing. This paper will explore the basis for these misconceptions, provide general insight into the current understanding of valve packing and demonstrate how with this new understanding and current valve diagnostic equipment the key aspects required to develop an effective, quality valve packing program fit together. The cost and operational benefits provided by this approach can be significant impact by the: elimination of periodic valve repacking, reduction of maintenance costs, benefits of leak-free valve operation, justification for reduced Post Maintenance Test Requirements, reduced radiation exposure, improved plant appearance.

  13. Soft valves in plants

    Science.gov (United States)

    Park, Keunhwan; Tixier, Aude; Christensen, Anneline; Arnbjerg-Nielsen, Sif; Zwieniecki, Maciej; Jensen, Kaare

    2017-11-01

    Water and minerals flow from plant roots to leaves in the xylem, an interconnected network of vascular conduits that spans the full length of the organism. When a plant is subjected to drought stress, air pockets can spread inside the xylem, threatening the survival of the plant. Many plants prevent propagation of air by using hydrophobic nano-membranes in the ``pit'' pores that link adjacent xylem cells. This adds considerable resistance to flow. Interestingly, torus-margo pit pores in conifers are open and offer less resistance. To prevent propagation of air, conifers use a soft gating mechanism, which relies on hydrodynamic interactions between the xylem liquid and the elastic pit. However, it is unknown exactly how it is able to combine the seemingly antagonist functions of high permeability and resistance to propagation of air. We conduct experiments on biomimetic pores to elucidate the flow regulation mechanism. The design of plant valves is compared to other natural systems and optimal strategies are discussed. This work was supported by a research Grant (13166) from VILLUM FONDEN.

  14. Annular flow diverter valve

    International Nuclear Information System (INIS)

    Rider, R.L.

    1980-01-01

    A valve is described for diverting flow from the center of two concentric tubes to the annulus between the tubes or, operating in the reverse direction, for mixing fluids from concentric tubes into a common tube and for controlling the volume ratio of said flow. It consists of a toroidal baffle disposed in sliding engagement with the interior of the inner tube downstream of a plurality of ports in the inner tube, a plurality of gates in sliding engagement with the interior of the inner tube attached to the baffle for movement therewith, a servomotor having a bullet-shaped plug on the downstream end thereof, and drive rods connecting the servomotor to the toroidal baffle. The sevomotor is adapted to move the baffle into mating engagement with the bullet-shaped plug and simultaneously move the gates away from the ports in the inner tube and to move the baffle away from the bullet-shaped plug and simultaneously move the gates to cover the ports in the inner tube

  15. Autopsy after transcatheter aortic valve implantation

    DEFF Research Database (Denmark)

    van Kesteren, F; Wiegerinck, E M A; Rizzo, S

    2017-01-01

    Autopsy after transcatheter aortic valve implantation (TAVI) is a new field of interest in cardiovascular pathology. To identify the cause of death, it is important to be familiar with specific findings related to the time interval between the procedure and death. We aimed to provide an overview...... different cause of death as was clinically determined. Autopsy on patients who underwent TAVI reveals specific patterns of cardiovascular pathology that clearly relate to the time interval between TAVI and death and significantly adds to the clinical diagnosis. Our data support the role of autopsy including...

  16. Ahmed glaucoma valve implant: surgical technique and complications.

    Science.gov (United States)

    Riva, Ivano; Roberti, Gloria; Oddone, Francesco; Konstas, Anastasios Gp; Quaranta, Luciano

    2017-01-01

    Implantation of Ahmed glaucoma valve is an effective surgical technique to reduce intraocular pressure in patients affected with glaucoma. While in the past, the use of this device was reserved to glaucoma refractory to multiple filtration surgical procedures, up-to-date mounting experience has encouraged its use also as a primary surgery for selected cases. Implantation of Ahmed glaucoma valve can be challenging for the surgeon, especially in patients who already underwent previous multiple surgeries. Several tips have to be acquired by the surgeon, and a long learning curve is always needed. Although the valve mechanism embedded in the Ahmed glaucoma valve decreases the risk of postoperative hypotony-related complications, it does not avoid the need of a careful follow-up. Complications related to this type of surgery include early and late postoperative hypotony, excessive capsule fibrosis around the plate, erosion of the tube or plate edge, and very rarely infection. The aim of this review is to describe surgical technique for Ahmed glaucoma valve implantation and to report related complications.

  17. LOFT pressurizer safety: relief valve reliability

    International Nuclear Information System (INIS)

    Brown, E.S.

    1978-01-01

    The LOFT pressurizer self-actuating safety-relief valves are constructed to the present state-of-the-art and should have reliability equivalent to the valves in use on PWR plants in the U.S. There have been no NRC incident reports on valve failures to lift that would challenge the Technical Specification Safety Limit. Fourteen valves have been reported as lifting a few percentage points outside the +-1% Tech. Spec. surveillance tolerance (9 valves tested over and 5 valves tested under specification). There have been no incident reports on failures to reseat. The LOFT surveillance program for assuring reliability is equivalent to nuclear industry practice

  18. Valve system incorporating single failure protection logic

    Science.gov (United States)

    Ryan, Rodger; Timmerman, Walter J. H.

    1980-01-01

    A valve system incorporating single failure protective logic. The system consists of a valve combination or composite valve which allows actuation or de-actuation of a device such as a hydraulic cylinder or other mechanism, integral with or separate from the valve assembly, by means of three independent input signals combined in a function commonly known as two-out-of-three logic. Using the input signals as independent and redundant actuation/de-actuation signals, a single signal failure, or failure of the corresponding valve or valve set, will neither prevent the desired action, nor cause the undesired action of the mechanism.

  19. LOFT pressurizer safety: relief valve reliability

    Energy Technology Data Exchange (ETDEWEB)

    Brown, E.S.

    1978-01-18

    The LOFT pressurizer self-actuating safety-relief valves are constructed to the present state-of-the-art and should have reliability equivalent to the valves in use on PWR plants in the U.S. There have been no NRC incident reports on valve failures to lift that would challenge the Technical Specification Safety Limit. Fourteen valves have been reported as lifting a few percentage points outside the +-1% Tech. Spec. surveillance tolerance (9 valves tested over and 5 valves tested under specification). There have been no incident reports on failures to reseat. The LOFT surveillance program for assuring reliability is equivalent to nuclear industry practice.

  20. Depth profiling of fluorine-doped diamond-like carbon (F-DLC) film: Localized fluorine in the top-most thin layer can enhance the non-thrombogenic properties of F-DLC

    Energy Technology Data Exchange (ETDEWEB)

    Hasebe, Terumitsu [Center for Science of Environment, Resources and Energy, Keio University Faculty of Science and Technology, 3-14-1 Hiyoshi, Kohoku-ku, Yokohama, Kanagawa 223-8522 (Japan); Department of Radiology, Tachikawa Hospital, 4-2-22, Nishiki-cho, Tachikawa, Tokyo 190-8531 (Japan)], E-mail: teru_hasebe@hotmail.com; Nagashima, So [Center for Science of Environment, Resources and Energy, Keio University Faculty of Science and Technology, 3-14-1 Hiyoshi, Kohoku-ku, Yokohama, Kanagawa 223-8522 (Japan); Kamijo, Aki [Department of Transfusion Medicine, the University of Tokyo Hospital, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8655 (Japan); Yoshimura, Taichi; Ishimaru, Tetsuya; Yoshimoto, Yukihiro; Yohena, Satoshi; Kodama, Hideyuki; Hotta, Atsushi [Center for Science of Environment, Resources and Energy, Keio University Faculty of Science and Technology, 3-14-1 Hiyoshi, Kohoku-ku, Yokohama, Kanagawa 223-8522 (Japan); Takahashi, Koki [Department of Transfusion Medicine, the University of Tokyo Hospital, 7-3-1 Hongo, Bunkyo-ku, Tokyo 113-8655 (Japan); Suzuki, Tetsuya [Center for Science of Environment, Resources and Energy, Keio University Faculty of Science and Technology, 3-14-1 Hiyoshi, Kohoku-ku, Yokohama, Kanagawa 223-8522 (Japan)

    2007-12-03

    Fluorine-doped diamond-like carbon (F-DLC) has recently drawn a great deal of attention as a more non-thrombogenic coating than conventional DLC for blood-contacting medical devices. We conducted quantitative depth profiling of F-DLC film by X-ray photoelectron spectroscopy (XPS) in order to elucidate the effects of fluorine and fluorine distribution in F-DLC film in connection with the prevention of surface blood adhesion. F-DLC films were prepared on silicon substrates using the radio frequency plasma enhanced chemical vapor deposition method, and the thickness of films was {approx} 50 nm. 50-nm-thick F-DLC film samples were etched at 10-nm thickness intervals using argon plasma, and each surface was examined by XPS. Thereafter, each etched film layer was incubated with platelet-rich plasma isolated from human whole blood, and the platelet-covered area per unit area was evaluated for each surface. XPS spectra showed the localization of doped fluorine in the top-most thin layer of the film. Platelet-covered areas represented progressively larger portions of the surfaces of deeper etched layers, corresponding to the decreasing fluorine content in such sample surfaces. These results indicate that the localized fluorine in the top-most thin layer is one of the key factors in the promotion of the non-thrombogenicity of F-DLC film.

  1. MANAGEMENT OF FAILED MITRAL VALVE REPLACEMENT. THE DURBAN EXPERIENCE.

    Science.gov (United States)

    Kistan, D; Booysen, M; Alexander, G; Madiba, T E

    2017-06-01

    Mitral valve replacement is the procedure of choice in patients with severe mitral valve disease. However, these patients are surviving longer and are thus at an increased risk of prosthesis failure or valve-related complications. Study setting: Inkosi Albert Luthuli Central Hospital, a tertiary referral Hospital in Durban. Study population: All patients undergoing redo mechanical mitral valve replacement surgery between January 2005 and December 2014. Study design: Retrospective analysis of patients undergoing redo mitral valve replacement. Patients were identified from theatre record books, their files were electronically accessed and pertinent information extracted onto a data capture sheet. Information documented included demographics, duration to failure, INR, Albumin, HIV status, clinical findings and outcome. The data was stored on an Excel datasheet. Fifty-eight patients were documented (mean age 32 ± 15.81 years; M:F 1:3). Ten patients (17%) were HIV positive (median CD4 count 478). Mean duration between first surgery and redo was 8.8 years. Thirty-five patients (60%) had no co-morbidities. Presenting features at redo surgery were congestive cardiac failure (27), chest pain (11) and palpitations (17). Mean preoperative Ejection Fraction was 51.65 %. Twenty-nine patients (55%) had emergency redo surgery. Twenty-two patients (75%) had acute prosthetic valve thrombosis. Thirty-two patients had tricuspid regurgitation. Original pathology was documented in 23 patients (40%) as Rheumatic valve disease. Prosthetic valve thrombosis was documented in 31 patients (54%). The most commonly used valve was the On-X. Mean presenting INR was 1.96 + 1.2 and mean presenting serum albumin was 36.7 + 7.8 g/l. Forty-one patients (71%) were found to be compliant to Warfarin therapy prior to redo surgery. Mean ICU stay was 6 +9 days. Two patients died postoperatively. Mean followup was 32 + 26.6 months. Twelve patients (20.7%) developed postoperative complications. Patients

  2. Risk-adjusted survival after tissue versus mechanical aortic valve replacement: a 23-year assessment.

    Science.gov (United States)

    Gaca, Jeffrey G; Clare, Robert M; Rankin, J Scott; Daneshmand, Mani A; Milano, Carmelo A; Hughes, G Chad; Wolfe, Walter G; Glower, Donald D; Smith, Peter K

    2013-11-01

    characteristics across the spectrum of patient risk. Further studies of the relative outcomes of mechanical versus tissue valves across the spectrum of patient age seem indicated.

  3. Two-dimensional FSI simulation of closing dynamics of a tilting disc mechanical heart valve.

    Science.gov (United States)

    Govindarajan, V; Udaykumar, H S; Herbertson, L H; Deutsch, S; Manning, K B; Chandran, K B

    2010-03-01

    The fluid dynamics during valve closure resulting in high shear flows and large residence times of particles has been implicated in platelet activation and thrombus formation in mechanical heart valves. Our previous studies with bi-leaflet valves have shown that large shear stresses induced in the gap between the leaflet edge and the valve housing results in relatively high platelet activation levels whereas flow between the leaflets results in shed vortices not conducive to platelet damage. In this study we compare the result of closing dynamics of a tilting disc valve with that of a bi-leaflet valve. The two-dimensional fluid-structure interaction analysis of a tilting disc valve closure mechanics is performed with a fixed grid Cartesian mesh flow solver with local mesh refinement, and a Lagrangian particle dynamic analysis for computation of potential for platelet activation. Throughout the simulation the flow remains in the laminar regime and the flow through the gap width is marked by the development of a shear layer which separates from the leaflet downstream of the valve. Zones of re-circulation are observed in the gap between the leaflet edge and the valve housing on the major orifice region of the tilting disc valve and are seen to be migrating towards the minor orifice region. Jet flow is observed at the minor orifice region and a vortex is formed which sheds in the direction of fluid motion as observed in experiments using PIV measurements. The activation parameter computed for the tilting disc valve, at the time of closure was found to be 2.7 times greater than that of the bi-leaflet mechanical valve and was found to be in the vicinity of the minor orifice region mainly due to the migration of vortical structures from the major to the minor orifice region during the leaflet rebound of the closing phase.

  4. Butterfly valve of all rubber lining type

    International Nuclear Information System (INIS)

    Shimada, Shosaku; Nakatsuma, Sumiya; Sasaki, Iwao; Aoki, Naoshi.

    1982-01-01

    The valves used for the circulating water pipes for condensers in nuclear and thermal power stations have become large with the increase of power output, and their specifications have become strict. The materials for the valves change from cast iron to steel plate construction. To cope with sea water corrosion, rubber lining has been applied to the internal surfaces of valve boxes, and the build-up welding of stainless steel has been made on the edges of valves. However, recently it is desired to develop butterfly valves, of which the whole valve disks are lined with hard rubber. For the purpose of confirming the performance of large bore valves, a 2600 mm bore butterfly valve of all rubber lining type was used, and the opening and closing test of 1100 times was carried out by applying thermal cycle and pressure difference and using artifical sea water. Also the bending test of hard rubber lining was performed with test pieces. Thus, it was confirmed that the butterfly valves of all rubber lining type have the performance exceeding that of the valves with build-up welding. The course of development of the valves of all rubber lining type, the construction and the items of confirmation by tests of these valves, and the tests of the valve and the hard rubber lining described above are reported. (Kako, I.)

  5. Door valve for fuel handling path

    International Nuclear Information System (INIS)

    Makishima, Katsuhiko.

    1969-01-01

    A door valve is provided which seals cover gas from a liquid metal cooled reactor without leakage therefrom. A threaded shaft is screwed into a heavy box press which is packed with lead. The shaft is adapted to be rotated by an electric motor or a manually operated wheel which is disposed outside of the door valve. A valve plate is suspended from the box press by four guide wheels mounted thereon. The guide wheels are fitted into inclined guide grooves formed at the valve plate and into grooved formed in the inner wall of a valve casing. A locking ball is provided at each side of the valve plate. In operation the shaft rotates and travels to permit the box press and the valve plate to move into the door valve casing, thus releasing the locking balls. The valve plate does not contact the bottom of the casing. When the box press reaches the home position, the valve plate is carried on the valve opening, and the box press presses the valve plate to increase the tightness. The valve plate does not suffer wear as it does not slide over other parts. (Yamaguchi, T.)

  6. Sequential transcatheter aortic valve implantation due to valve dislodgement - a Portico valve implanted over a CoreValve bioprosthesis.

    Science.gov (United States)

    Campante Teles, Rui; Costa, Cátia; Almeida, Manuel; Brito, João; Sondergaard, Lars; Neves, José P; Abecasis, João; M Gabriel, Henrique

    2017-03-01

    Transcatheter aortic valve implantation (TAVI) has become an important treatment in high surgical risk patients with severe aortic stenosis (AS), whose complications need to be managed promptly. The authors report the case of an 86-year-old woman presenting with severe symptomatic AS, rejected for surgery due to advanced age and comorbidities. The patient underwent a first TAVI, with implantation of a Medtronic CoreValve ® , which became dislodged and migrated to the ascending aorta. Due to the previous balloon valvuloplasty, the patient's AS became moderate, and her symptoms improved. After several months, she required another intervention, performed with a St. Jude Portico ® repositionable self-expanding transcatheter aortic valve. There was a good clinical response that was maintained at one-year follow-up. The use of a self-expanding transcatheter bioprosthesis with repositioning features is a solution in cases of valve dislocation to avoid suboptimal positioning of a second implant, especially when the two valves have to be positioned overlapping or partially overlapping each other. Copyright © 2017 Sociedade Portuguesa de Cardiologia. Publicado por Elsevier España, S.L.U. All rights reserved.

  7. Cavitation noise from butterfly valves

    International Nuclear Information System (INIS)

    Rahmeyer, W.J.

    1982-01-01

    Cavitation in valves can produce levels of intense noise. It is possible to mathematically express a limit for a design level of cavitation noise in terms of the cavitation parameter sigma. Using the cavitation parameter or limit, it is then possible to calculate the flow conditions at which a design level of cavitation noise will occur. However, the intensity of cavitation increases with the upstream pressure and valve size at a constant sigma. Therefore, it is necessary to derive equations to correct or scale the cavitation limit for the effects of different upstream pressures and valve sizes. The following paper discusses and presents experimental data for the caviation noise limit as well as the cavitation limits of incipient, critical, incipient damage, and choking cavitation for butterfly valves. The main emphasis is on the design limit of caviation noise, and a noise level of 85 decibels was selected as the noise limit. Tables of data and scaling exponents are included for applying the design limits for the effects of upstream pressure and valve size. (orig.)

  8. Early results after implantation of a new geometric annuloplasty ring for aortic valve repair.

    Science.gov (United States)

    Mazzitelli, Domenico; Nöbauer, Christian; Rankin, J Scott; Badiu, Catalin C; Krane, Markus; Crooke, Philip S; Cohn, William E; Opitz, Anke; Schreiber, Christian; Lange, Rüdiger

    2013-01-01

    Aortic valve repair is associated with fewer long-term valve-related complications as compared with valve replacement, and repair is being performed increasingly. A current problem is the lack of a geometric annuloplasty ring to facilitate reconstruction. This paper describes the first clinical application of such a device designed to permanently restore physiologic annular size and geometry during aortic valve repair. Based on mathematical studies of human cadaver valves, as well as computed tomography angiographic analyses of awake patients with normal valves, a three-dimensional annuloplasty ring has been developed, consisting of low-profile, one-piece titanium construction and Dacron cloth covering. The ring design incorporates 2:3 elliptical base geometry and 10-degree outwardly flaring subcommissural posts. Appropriately sized rings were implanted in 5 patients with severe aortic insufficiency due to annular dilation and anatomic leaflet defects. The rings restored annular geometry and facilitated leaflet repairs in all patients. Each recovered excellent valve function with minimal residual leak. All patients convalesced uneventfully, were discharged within 7 days after surgery, and continue with stable valve function as long as 6 months after implantation. Initial clinical application of a geometric aortic annuloplasty ring was associated with excellent device performance and perhaps better repairs. Further clinical series and patient follow-up should identify potential benefits of the device, including improved applicability and stability of aortic valve repair. Copyright © 2013 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  9. Overview of Prevention for Water Hammer by Check Valve Action in Nuclear Reactor

    International Nuclear Information System (INIS)

    Kim, Dayong; Yoon, Hyungi; Seo, Kyoungwoo; Kim, Seonhoon

    2016-01-01

    Water hammer can cause serious damage to pumping system and unexpected system pressure rise in the pipeline. In nuclear reactor, water hammer can influence on the integrity of safety related system. Water hammer in nuclear reactor have been caused by voiding in normally water-filled lines, steam condensation line containing both steam and water, as well as by rapid check valve action. Therefore, this study focuses on the water hammer by check valve among the sources of water hammer occurrence and suggests proper methodology for check valve type selection against water hammer. This study focuses on the water hammer by check valve action among the sources of water hammer occurrence and suggests proper methodology for check valve type selection against water hammer. If the inadvertent pump trip or pipe rupture in high velocity and pressure pipe is predicted, the fast response check valve such as tiled disc, dual disc and nozzle check valve should be installed in the system. If the inadvertent pump trip or pipe rupture in very high velocity and pressure pipe and excessively large revered flow velocity are predicted, the very slowly closing check valve such as controlled closure check valve should be installed in the system

  10. Overview of Prevention for Water Hammer by Check Valve Action in Nuclear Reactor

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Dayong; Yoon, Hyungi; Seo, Kyoungwoo; Kim, Seonhoon [Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)

    2016-10-15

    Water hammer can cause serious damage to pumping system and unexpected system pressure rise in the pipeline. In nuclear reactor, water hammer can influence on the integrity of safety related system. Water hammer in nuclear reactor have been caused by voiding in normally water-filled lines, steam condensation line containing both steam and water, as well as by rapid check valve action. Therefore, this study focuses on the water hammer by check valve among the sources of water hammer occurrence and suggests proper methodology for check valve type selection against water hammer. This study focuses on the water hammer by check valve action among the sources of water hammer occurrence and suggests proper methodology for check valve type selection against water hammer. If the inadvertent pump trip or pipe rupture in high velocity and pressure pipe is predicted, the fast response check valve such as tiled disc, dual disc and nozzle check valve should be installed in the system. If the inadvertent pump trip or pipe rupture in very high velocity and pressure pipe and excessively large revered flow velocity are predicted, the very slowly closing check valve such as controlled closure check valve should be installed in the system.

  11. Increased amount of atrial fibrosis in patients with atrial fibrillation secondary to mitral valve disease

    NARCIS (Netherlands)

    Geuzebroek, Guillaume S. C.; van Amersfoorth, Shirley C. M.; Hoogendijk, Mark G.; Kelder, Johannes C.; van Hemel, Norbert M.; de Bakker, Jacques M. T.; Coronel, Ruben

    2012-01-01

    Objective: Atrial fibrosis is related to atrial fibrillation but may differ in patients with mitral valve disease or lone atrial fibrillation. Therefore, we studied atrial fibrosis in patients with atrial fibrillation + mitral valve disease or with lone atrial fibrillation and compared it with

  12. Structural valve deterioration in the Mitroflow biological heart valve prosthesis

    DEFF Research Database (Denmark)

    Issa, Issa Farah; Poulsen, Steen Hvitfeldt; Waziri, Farhad

    2018-01-01

    OBJECTIVES: Concern has been raised regarding the long-term durability of the Mitroflow biological heart valve prosthesis. Our aim was to assess the incidence of structural valve degeneration (SVD) for the Mitroflow bioprosthesis in a nationwide study in Denmark including all patients alive......: A total of 173 patients were diagnosed with SVD by echocardiography. Of these, 64 (11%) patients had severe SVD and 109 (19%) patients moderate SVD. Severe SVD was associated with the age of the prosthesis and small prosthesis size [Size 21: hazard ratio (95% confidence interval, CI) 2.72 (0.97-8.56), P...

  13. Influence of surgical implantation angle of left ventricular assist device outflow graft and management of aortic valve opening on the risk of stroke in heart failure patients

    Science.gov (United States)

    Chivukula, V. Keshav; McGah, Patrick; Prisco, Anthony; Beckman, Jennifer; Mokadam, Nanush; Mahr, Claudius; Aliseda, Alberto

    2016-11-01

    Flow in the aortic vasculature may impact stroke risk in patients with left ventricular assist devices (LVAD) due to severely altered hemodynamics. Patient-specific 3D models of the aortic arch and great vessels were created with an LVAD outflow graft at 45, 60 and 90° from centerline of the ascending aorta, in order to understand the effect of surgical placement on hemodynamics and thrombotic risk. Intermittent aortic valve opening (once every five cardiac cycles) was simulated and the impact of this residual native output investigated for the potential to wash out stagnant flow in the aortic root region. Unsteady CFD simulations with patient-specific boundary conditions were performed. Particle tracking for 10 cardiac cycles was used to determine platelet residence times and shear stress histories. Thrombosis risk was assessed by a combination of Eulerian and Lagrangian metrics and a newly developed thrombogenic potential metric. Results show a strong influence of LVAD outflow graft angle on hemodynamics in the ascending aorta and consequently on stroke risk, with a highly positive impact of aortic valve opening, even at low frequencies. Optimization of LVAD implantation and management strategies based on patient-specific simulations to minimize stroke risk will be presented

  14. Long-Term Results of Mitral Valve Repair

    Directory of Open Access Journals (Sweden)

    Francisco Diniz Affonso da Costa

    Full Text Available Abstract Introduction: Current guidelines state that patients with severe mitral regurgitation should be treated in reference centers with a high reparability rate, low mortality rate, and durable results. Objective: To analyze our global experience with the treatment of organic mitral regurgitation from various etiologies operated in a single center. Methods: We evaluated all surgically treated patients with organic mitral regurgitation from 2004-2017. Patients were evaluated clinically and by echocardiography every year. We determined early and late survival rates, valve related events and freedom from recurrent mitral regurgitation and tricuspid regurgitation. Valve failure was defined as any mitral regurgitation ≥ moderate degree or the need for reoperation for any reason. Results: Out of 133 patients with organic mitral regurgitation, 125 (93.9% were submitted to valve repair. Mean age was 57±15 years and 52 patients were males. The most common etiologies were degenerative disease (73 patients and rheumatic disease (34 patients. Early mortality was 2.4% and late survival was 84.3% at 10 years, which are similar to the age- and gender-matched general population. Only two patients developed severe mitral regurgitation, and both were reoperated (95.6% at 10 years. Freedom from mitral valve failure was 84.5% at 10 years, with no difference between degenerative and rheumatic valves. Overall, late ≥ moderate tricuspid regurgitation was present in 34% of the patients, being more common in the rheumatic ones. The use of tricuspid annuloplasty abolished this complication. Conclusion: We have demonstrated that mitral regurgitation due to organic mitral valve disease from various etiologies can be surgically treated with a high repair rate, low early mortality and long-term survival that are comparable to the matched general population. Concomitant treatment of atrial fibrillation and tricuspid valve may be important adjuncts to optimize long

  15. Outcomes in nonagenarians after heart valve replacement operation.

    Science.gov (United States)

    Edwards, Maria-Benedicta; Taylor, Kenneth M

    2003-03-01

    Changes in the age profile of the United Kingdom population and improvements in preoperative and postoperative care have resulted in increasing numbers of very elderly patients undergoing heart valve replacement (HVR) operations. Although HVR operations in nonagenarians are relatively uncommon, the demand for cardiac operations in this age group may increase over time. Outcomes after HVR operations in nonagenarians have not been well described yet. Therefore, the aim of this study was to determine outcomes in terms of early mortality and long-term survival in 35 nonagenarians after HVR operation. Data from the United Kingdom Heart Valve Registry were analyzed and nonagenarian patients were identified. Additional analyzed data include gender, valve position, valve type, valve size, operative priority, follow-up time, and date and cause of death. Kaplan-Meier actuarial curves were calculated to determine accurate 30-day mortality and long-term survival. On average five HVR operations are performed annually in the United Kingdom in nonagenarians with equal numbers of males and females. Aortic valve replacement with a bioprosthetic valve was the most common operation and 86% were elective admissions. Fourteen patients died within the review period; mean time to death was 402 days. Overall 30-day mortality was 17%, which was higher for males compared with females; females also displayed better long-term survival. HVR operations in nonagenarians carry a significantly higher risk of early mortality and reduced long-term survival. Despite increases in the age profile of the population, elective HVR operation with patients aged 90 years or older is likely to remain an infrequent surgical procedure reserved for very carefully selected patients.

  16. Bicuspid Aortic Valve Disease: A Comprehensive Review

    OpenAIRE

    Mordi, Ify; Tzemos, Nikolaos

    2012-01-01

    Bicuspid aortic valve is the commonest congenital cardiac abnormality in the general population. This paper article will discuss our current knowledge of the anatomy, pathophysiology, genetics, and clinical aspects of bicuspid aortic valve disease.

  17. Comparative study between CardiaMed valves (freely floating valve leaflets versus St. Jude Medical (fixed valve leaflets in mitral valve replacement surgery

    Directory of Open Access Journals (Sweden)

    Mostafa Ahmed

    2017-09-01

    Conclusions: CardiaMed freely floating leaflet prostheses showed good hemodynamic characteristics. The prosthesis adequately corrects hemodynamics and is safe and no worse than the St. Jude Medical valve in the mitral valve position.

  18. Prosthetic Mitral Valve Leaflet Escape

    Science.gov (United States)

    Kim, Darae; Hun, Sin Sang; Cho, In-Jeong; Shim, Chi-Young; Ha, Jong-Won; Chung, Namsik; Ju, Hyun Chul; Sohn, Jang Won

    2013-01-01

    Leaflet escape of prosthetic valve is rare but potentially life threatening. It is essential to make timely diagnosis in order to avoid mortality. Transesophageal echocardiography and cinefluoroscopy is usually diagnostic and the location of the missing leaflet can be identified by computed tomography (CT). Emergent surgical correction is mandatory. We report a case of fractured escape of Edward-Duromedics mitral valve 27 years after the surgery. The patient presented with symptoms of acute decompensated heart failure and cardiogenic shock. She was instantly intubated and mechanically ventilated. After prompt evaluation including transthoracic echocardiography and CT, the escape of the leaflet was confirmed. The patient underwent emergent surgery for replacement of the damaged prosthetic valves immediately. Eleven days after the surgery, the dislodged leaflet in iliac artery was removed safely and the patient recovered well. PMID:23837121

  19. Active combustion flow modulation valve

    Science.gov (United States)

    Hensel, John Peter; Black, Nathaniel; Thorton, Jimmy Dean; Vipperman, Jeffrey Stuart; Lambeth, David N; Clark, William W

    2013-09-24

    A flow modulation valve has a slidably translating hollow armature with at least one energizable coil wound around and fixably attached to the hollow armature. The energizable coil or coils are influenced by at least one permanent magnet surrounding the hollow armature and supported by an outer casing. Lorentz forces on the energizable coils which are translated to the hollow armature, increase or decrease the flow area to provide flow throttling action. The extent of hollow armature translation depends on the value of current supplied and the direction of translation depends on the direction of current flow. The compact nature of the flow modulation valve combined with the high forces afforded by the actuator design provide a flow modulation valve which is highly responsive to high-rate input control signals.

  20. Statins for aortic valve stenosis

    Directory of Open Access Journals (Sweden)

    Luciana Thiago

    Full Text Available ABSTRACT BACKGROUND: Aortic valve stenosis is the most common type of valvular heart disease in the USA and Europe. Aortic valve stenosis is considered similar to atherosclerotic disease. Some studies have evaluated statins for aortic valve stenosis. OBJECTIVES: To evaluate the effectiveness and safety of statins in aortic valve stenosis. METHODS: Search methods: We searched the Cochrane Central Register of Controlled Trials (CENTRAL, MEDLINE, Embase, LILACS - IBECS, Web of Science and CINAHL Plus. These databases were searched from their inception to 24 November 2015. We also searched trials in registers for ongoing trials. We used no language restrictions. Selection criteria: Randomized controlled clinical trials (RCTs comparing statins alone or in association with other systemic drugs to reduce cholesterol levels versus placebo or usual care. Data collection and analysis: Primary outcomes were severity of aortic valve stenosis (evaluated by echocardiographic criteria: mean pressure gradient, valve area and aortic jet velocity, freedom from valve replacement and death from cardiovascular cause. Secondary outcomes were hospitalization for any reason, overall mortality, adverse events and patient quality of life. Two review authors independently selected trials for inclusion, extracted data and assessed the risk of bias. The GRADE methodology was employed to assess the quality of result findings and the GRADE profiler (GRADEPRO was used to import data from Review Manager 5.3 to create a 'Summary of findings' table. MAIN RESULTS: We included four RCTs with 2360 participants comparing statins (1185 participants with placebo (1175 participants. We found low-quality evidence for our primary outcome of severity of aortic valve stenosis, evaluated by mean pressure gradient (mean difference (MD -0.54, 95% confidence interval (CI -1.88 to 0.80; participants = 1935; studies = 2, valve area (MD -0.07, 95% CI -0.28 to 0.14; participants = 127; studies = 2

  1. Statins for aortic valve stenosis.

    Science.gov (United States)

    Thiago, Luciana; Tsuji, Selma Rumiko; Nyong, Jonathan; Puga, Maria Eduarda Dos Santos; Góis, Aécio Flávio Teixeira de; Macedo, Cristiane Rufino; Valente, Orsine; Atallah, Álvaro Nagib

    2016-01-01

    Aortic valve stenosis is the most common type of valvular heart disease in the USA and Europe. Aortic valve stenosis is considered similar to atherosclerotic disease. Some studies have evaluated statins for aortic valve stenosis. To evaluate the effectiveness and safety of statins in aortic valve stenosis. Search methods: We searched the Cochrane Central Register of Controlled Trials (CENTRAL), MEDLINE, Embase, LILACS - IBECS, Web of Science and CINAHL Plus. These databases were searched from their inception to 24 November 2015. We also searched trials in registers for ongoing trials. We used no language restrictions.Selection criteria: Randomized controlled clinical trials (RCTs) comparing statins alone or in association with other systemic drugs to reduce cholesterol levels versus placebo or usual care. Data collection and analysis: Primary outcomes were severity of aortic valve stenosis (evaluated by echocardiographic criteria: mean pressure gradient, valve area and aortic jet velocity), freedom from valve replacement and death from cardiovascular cause. Secondary outcomes were hospitalization for any reason, overall mortality, adverse events and patient quality of life.Two review authors independently selected trials for inclusion, extracted data and assessed the risk of bias. The GRADE methodology was employed to assess the quality of result findings and the GRADE profiler (GRADEPRO) was used to import data from Review Manager 5.3 to create a 'Summary of findings' table. We included four RCTs with 2360 participants comparing statins (1185 participants) with placebo (1175 participants). We found low-quality evidence for our primary outcome of severity of aortic valve stenosis, evaluated by mean pressure gradient (mean difference (MD) -0.54, 95% confidence interval (CI) -1.88 to 0.80; participants = 1935; studies = 2), valve area (MD -0.07, 95% CI -0.28 to 0.14; participants = 127; studies = 2), and aortic jet velocity (MD -0.06, 95% CI -0.26 to 0

  2. Evaluation of mispositioned ECCS valves

    International Nuclear Information System (INIS)

    Hill, R.A.; O'Brien, J.F.; McIntire, D.C.; Barlow, R.T.

    1977-09-01

    In October of 1975, Westinghouse submitted NS-CE-787, dated October 17, 1975, to the Nuclear Regulatory Commission (NRC) and entered into discussions with them concerning the spurious movement of certain motor-operated valves (MOV's) in the Emergency Core Cooling System (ECCS) to a position defeating the ECCS function at a time when this function is required. On November 25, 1975, the discussion turned to the possible movement of a manually controlled, motor-operated valve due to a fault in its electrical circuitry and the NRC staff expressed concerns about other possible failure modes that might lead to such a valve movement. The NRC meeting minutes document these concerns. This report is an item-by-item response to the concerns expressed by the NRC staff at that meeting and incorporates the original electrical fault analysis

  3. Pannus Formation Leads to Valve Malfunction in the Tricuspid Position 19 Years after Triple Valve Replacement.

    Science.gov (United States)

    Alskaf, Ebraham; McConkey, Hannah; Laskar, Nabila; Kardos, Attila

    2016-06-20

    The Medtronic ATS Open Pivot mechanical valve has been successfully used in heart valve surgery for more than two decades. We present the case of a patient who, 19 years following a tricuspid valve replacement with an ATS prosthesis as part of a triple valve operation following infective endocarditis, developed severe tricuspid regurgitation due to pannus formation.

  4. The nordic aortic valve intervention (NOTION) trial comparing transcatheter versus surgical valve implantation

    DEFF Research Database (Denmark)

    Thyregod, Hans Gustav; Søndergaard, Lars; Ihlemann, Nikolaj

    2013-01-01

    Degenerative aortic valve (AV) stenosis is the most prevalent heart valve disease in the western world. Surgical aortic valve replacement (SAVR) has until recently been the standard of treatment for patients with severe AV stenosis. Whether transcatheter aortic valve implantation (TAVI) can...

  5. Early clinical outcome of aortic transcatheter valve-in-valve implantation in the Nordic countries

    DEFF Research Database (Denmark)

    Ihlberg, Leo; Nissen, Henrik Hoffmann; Nielsen, Niels Erik

    2013-01-01

    Transcatheter valve-in-valve implantation has emerged as an option, in addition to reoperative surgical aortic valve replacement, to treat failed biologic heart valve substitutes. However, the clinical experience with this approach is still limited. We report the comprehensive experience...

  6. Intro to Valve Guide Reconditioning. Automotive Mechanics. Valves. Instructor's Guide [and] Student Guide.

    Science.gov (United States)

    Horner, W.

    This instructional package, one in a series of individualized instructional units on tools and techniques for repairing worn valve guides in motor vehicles, provides practical experience for students in working on cylinder heads. Covered in the module are reaming valve guides that are oversized to match a new oversized valve, reaming valve guides…

  7. Infective Endocarditis of the Aortic Valve with Anterior Mitral Valve Leaflet Aneurysm

    NARCIS (Netherlands)

    Tomsic, Anton; Li, Wilson W. L.; van Paridon, Marieke; Bindraban, Navin R.; de Mol, Bas A. J. M.

    2016-01-01

    Mitral valve leaflet aneurysm is a rare and potentially devastating complication of aortic valve endocarditis. We report the case of a 48-year-old man who had endocarditis of the native aortic valve and a concomitant aneurysm of the anterior mitral valve leaflet. Severe mitral regurgitation occurred

  8. [Late complications following Björk-Shiley and St. Jude Medical heart valve replacement].

    Science.gov (United States)

    Horstkotte, D; Körfer, R; Budde, T; Haerten, K; Schulte, H D; Bircks, W; Loogen, F

    1983-05-01

    Valve-related complications after Björk-Shiley mitral (n = 475), aortic (n = 424), or mitral-aortic implantation (n = 119) were compared to complications after St. Jude mitral (n = 173), aortic (n = 152), and St. Jude mitral and aortic (n = 63) replacements. The 1,018 consecutive patients with Björk-Shiley valves had been operated upon between 1974 and 1982, those with St. Jude valves between 1978 and 1982. All patients were placed on anticoagulant therapy with phenprocoumon early after operation and no significant intergroup differences in the effectiveness of the anticoagulant therapy were found. At a comparable follow-up time of approximately 23 months, 24 major thromboembolic episodes were observed after Björk-Shiley mitral (BSM) and 3 after St. Jude mitral valve implantation (SJM), corresponding to a thromboembolic rate of 2.82/100 patient years with BSM and 0.93/100 patient years with SJM. After aortic valve replacements, 1.93 events in 100 patient years occurred after Björk-Shiley aortic (BSA) and 0.73 after St. Jude aortic implantation (SJA). In patients with double valve replacements, these rates were 3.2 (BSM + BSA) and 0.88 (SJM + SJA), respectively. The cerebral vessels were involved in 52% and the arteries of the extremities in 22% of these major events. Six Björk-Shiley prostheses had to be replaced because of valve thrombosis. The overall incidence of severe hemorrhagic complications was 2.94/100 patient years in BSM and 1.79 in SJM. After aortic valve replacement, we found rates of 1.80/100 patient years (BSA) and 2.57/100 patient years (SJA), respectively. Intravascular hemolysis no longer seems to be a significant clinical problem. However, indications of red cell damage after heart valve replacement were significantly greater in patients with perivalvular leakage, valve thrombosis, or dysfunction than in those with normally functioning prostheses. Reoperations were necessary because of valve thrombosis (0.46%), perivalvular leakage (2

  9. Valve-Sparing Root Replacement Compared With Composite Valve Graft Procedures in Patients With Aortic Root Dilation.

    Science.gov (United States)

    Ouzounian, Maral; Rao, Vivek; Manlhiot, Cedric; Abraham, Nachum; David, Carolyn; Feindel, Christopher M; David, Tirone E

    2016-10-25

    Although aortic valve-sparing (AVS) operations are established alternatives to composite valve graft (CVG) procedures for patients with aortic root aneurysms, comparative long-term outcomes are lacking. This study sought to compare the results of patients undergoing AVS procedures with those undergoing CVG operations. From 1990 to 2010, a total of 616 patients age Marfan syndrome and lower rates of bicuspid aortic valve than those undergoing bio-CVG or m-CVG procedures. In-hospital mortality (0.3%) and stroke rate (1.3%) were similar among groups. After adjusting for clinical covariates, both bio-CVG and m-CVG procedures were associated with increased long-term major adverse valve-related events compared with patients undergoing AVS (hazard ratio [HR]: 3.4, p = 0.005; and HR: 5.2, p valve-related complications when compared with bio-CVG and m-CVG. AVS is the treatment of choice for young patients with aortic root aneurysm and normal or near-normal aortic cusps. Copyright © 2016 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  10. Small sodium valve design and operating experience

    International Nuclear Information System (INIS)

    McGough, C.B.

    1974-01-01

    The United States Liquid Metal Fast Breeder Reactor program (LMFBR) includes an extensive program devoted to the development of small sodium valves. This program is now focused on the development and production of valves for the Fast Flux Test Facility (FFTF) now under construction near Richland, Washington. Other AEC support facilities, such as various test loops located at the Liquid Metal Engineering Center (LMEC), Los Angeles, California, and at the Hanford Engineering Development Laboratory (HEDL), Richland, Washington, also have significant requirements for small sodium valves, and valves similar in design to the FFTF valves are being supplied to these AEC laboratories for use in their critical test installations. A principal motivation for these valve programs, beyond the immediate need to provide high-reliability valves for FFTF and the support facilities, is the necessity to develop small valve technology for the Clinch River Breeder Reactor Plant (CRBRP). FFTF small sodium valve design and development experience will be directly applied to the CRBRP program. Various test programs have been, and are being, conducted to verify the performance and integrity of the FFTF valves, and to uncover any potential problems so that they can be corrected before the valves are placed in service in FFTF. The principal small sodium valve designs being utilized in current U.S. programs, the test and operational experience obtained to date on them, problems uncovered, and future development and testing efforts being planned are reviewed. The standards and requirements to which the valves are being designed and fabricated, the valve designs in current use, valve operators, test and operating experience, and future valve development plans are summarized. (U.S.)

  11. Promising results after percutaneous mitral valve repair

    DEFF Research Database (Denmark)

    Ihlemann, Nikolaj; Franzen, Olaf; Jørgensen, Erik

    2011-01-01

    Mitral valve regurgitation (MR) is the secondmost frequent valve disease in Europe. Untreated MR causes considerable morbidity and mortality. In the elderly, as many as half of these patients are denied surgery because of an estimated high surgical risk. Percutaneous mitral valve repair with the ...... with the MitraClip system resembles the Alfieristitch where a clip is used to connect the tip of the mitral valve leaflets....

  12. Infective endocarditis following percutaneous pulmonary valve replacement

    DEFF Research Database (Denmark)

    Cheung, Gary; Vejlstrup, Niels; Ihlemann, Nikolaj

    2013-01-01

    Infective endocarditis (IE) following percutaneous pulmonary valve replacement (PPVR) with the Melody valve is rarely reported. Furthermore, there are challenges in this diagnosis; especially echocardiographic evidence of vegetation within the prosthesis may be difficult.......Infective endocarditis (IE) following percutaneous pulmonary valve replacement (PPVR) with the Melody valve is rarely reported. Furthermore, there are challenges in this diagnosis; especially echocardiographic evidence of vegetation within the prosthesis may be difficult....

  13. Fast-acting valve actuator

    Science.gov (United States)

    Cho, Nakwon

    1980-01-01

    A fast-acting valve actuator utilizes a spring driven pneumatically loaded piston to drive a valve gate. Rapid exhaust of pressurized gas from the pneumatically loaded side of the piston facilitates an extremely rapid piston stroke. A flexible selector diaphragm opens and closes an exhaust port in response to pressure differentials created by energizing and de-energizing a solenoid which controls the pneumatic input to the actuator as well as selectively providing a venting action to one side of the selector diaphragm.

  14. Aortic valve replacement with the Biocor PSB stentless xenograft.

    Science.gov (United States)

    Bertolini, P; Luciani, G B; Vecchi, B; Pugliese, P; Mazzucco, A

    1998-08-01

    The midterm clinical results after aortic valve replacement with the Biocor PSB stentless xenograft on all patients operated between October 1992 and October 1996 were reviewed. One hundred six patients, aged 70+/-6 years, had aortic valve replacement for aortic stenosis (67%), regurgitation (11%), or both (22%). Associated procedures were done in 49 patients (46%), including coronary artery bypass in 30 patients, mitral valve repair/replacement in 16, and ascending aorta replacement in 5 patients. Aortic cross-clamp and cardiopulmonary bypass times were 96+/-24 and 129+/-31 minutes, respectively. There were 3 (3%) early deaths due to low output (2 patients) and cerebrovascular accident (1 patient). Follow-up of survivors ranged from 6 to 66 months (mean, 39+/-14 months). Survival was 94%+/-2% and 90%+/-3% at 1 and 5 years. There were 5 late deaths due to cardiac cause (2), cancer (2), and pulmonary embolism (1 patient). No patient had structural valve deterioration, whereas 100% and 95%+/-3% were free from valve-related events at 1 and 5 years. There were two reoperations due to narrowing of the left coronary ostium and endocarditis, with an actuarial freedom from reoperation of 99%+/-1% and 98+/-1% at 1 and 5 years, respectively. Functional results demonstrated a mean peak transprosthetic gradient of 16+/-12 mm Hg, with only 1 patient (1%) with a 55 mm Hg gradient. No cases of valve regurgitation greater than mild were recorded at follow-up. Assessment of New York Heart Association functional class demonstrated a significant improvement (2.9+/-0.6 versus 1.4+/-0.7; p=0.01). All patients were free from anticoagulation. Aortic valve replacement using the Biocor PSB stentless xenograft offers excellent midterm survival, negligible valve deterioration, and a very low rate of valve-related events, which are comparable to estimates reported with other models of stentless xenografts and currently available stented xenografts. Hemodynamic performance is favorable and

  15. Effects of the blockage ratio of a valve disk on loss coefficient in a butterfly valve

    International Nuclear Information System (INIS)

    Rho, Hyung Joon; Lee, Jee Keun; Choi, Hee Joo

    2008-01-01

    The loss coefficient of the butterfly valve which allows partial opening of the valve at closed position and is applicable to the small-sized pipe system with the diameter of 1 inch was measured for the variation of the valve disk blockage ratio. Two different types of the valve disk configuration to adjust the blockage ratio were considered. One was the solid type valve disk of which the diameter was changed into the smaller size rather than the pipe diameter, and the other was the perforate type valve disk on which some holes were perforated. The results from two types of valve disk were compared to identify their characteristics in the loss coefficient distributions. The loss coefficient and the controllable angle of the valve disk were decreased exponentially with the decrease of the blockage ratio. In addition, the perforate valve disk had the effect on the higher loss coefficient rather than the solid type valve disk

  16. 49 CFR 195.260 - Valves: Location.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 3 2010-10-01 2010-10-01 false Valves: Location. 195.260 Section 195.260... PIPELINE Construction § 195.260 Valves: Location. A valve must be installed at each of the following locations: (a) On the suction end and the discharge end of a pump station in a manner that permits isolation...

  17. Porcine Tricuspid Valve Anatomy and Human Compatibility

    DEFF Research Database (Denmark)

    Waziri, Farhad; Lyager Nielsen, Sten; Hasenkam, J. Michael

    2016-01-01

    before clinical use. The study aim was to evaluate and compare the tricuspid valve anatomy of porcine and human hearts. METHODS: The anatomy of the tricuspid valve and the surrounding structures that affect the valve during a cardiac cycle were examined in detail in 100 fresh and 19 formalin...

  18. Valve-sparing aortic root replacement†

    NARCIS (Netherlands)

    Koolbergen, David R.; Manshanden, Johan S. J.; Bouma, Berto J.; Blom, Nico A.; Mulder, Barbara J. M.; de Mol, Bas A. J. M.; Hazekamp, Mark G.

    2015-01-01

    To evaluate our results of valve-sparing aortic root replacement and associated (multiple) valve repair. From September 2003 to September 2013, 97 patients had valve-sparing aortic root replacement procedures. Patient records and preoperative, postoperative and recent echocardiograms were reviewed.

  19. Solving the problem of valve stem leakage

    International Nuclear Information System (INIS)

    Dixon, D.F.

    1976-01-01

    Engineering solutions to valve stem leakage, in systems carrying expensive heavy water under pressure, have progressed from changing packing brands (failure) to leak collection (partial success) to elimination of small packed valves and an improved valve packing strategy involving stable packing materials, live Belleville spring-loading of packing, and issuance of a detailed stuffing box specification (success). (E.C.B.)

  20. 49 CFR 229.109 - Safety valves.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 4 2010-10-01 2010-10-01 false Safety valves. 229.109 Section 229.109..., DEPARTMENT OF TRANSPORTATION RAILROAD LOCOMOTIVE SAFETY STANDARDS Safety Requirements Steam Generators § 229.109 Safety valves. Every steam generator shall be equipped with at least two safety valves that have a...

  1. Miniature piezo electric vacuum inlet valve

    Science.gov (United States)

    Keville, Robert F.; Dietrich, Daniel D.

    1998-03-24

    A miniature piezo electric vacuum inlet valve having a fast pulse rate and is battery operated with variable flow capability. The low power (piezo electric valves which require preloading of the crystal drive mechanism and 120 Vac, thus the valve of the present invention is smaller by a factor of three.

  2. Valve Corporation: Strategy Tipping Points and Thresholds

    OpenAIRE

    Teppo Felin

    2015-01-01

    Valve Corporation represents an intriguing case study of flat structure and self organization (Puranam & Håkonsson, 2015; Valve, 2012).  The structures and practices of Valve of course are not new. But the company provides an interesting experiment and illustration that powerfully highlights how organizational design can impact individual and collective behavior, strategy and performance.

  3. Valve Corporation: Strategy Tipping Points and Thresholds

    Directory of Open Access Journals (Sweden)

    Teppo Felin

    2015-06-01

    Full Text Available Valve Corporation represents an intriguing case study of flat structure and self organization (Puranam & Håkonsson, 2015; Valve, 2012.  The structures and practices of Valve of course are not new. But the company provides an interesting experiment and illustration that powerfully highlights how organizational design can impact individual and collective behavior, strategy and performance.

  4. Door valve for fuel handling path

    International Nuclear Information System (INIS)

    Makishima, Katsuhiko.

    1969-01-01

    A door valve is provided which seals cover gas from a liquid metal cooled reactor without leakage therefrom. A threaded shaft is screwed into a heavy box press which is packed with lead. The shaft is adapted to be rotated by an electric motor or a manually operated wheel which is disposed outside of the door valve. From the box press a valve plate is suspended by four linkage bars, one for each corner. Each linkage bar is provided with two wheels which are respectively mounted at the connections with the box press and the valve plate. The wheels are carried on the horizontal grooves formed in a door valve casing. In operation the shaft rotates and travels to permit the box press and the valve plate to move into the door valve casing while the valve plate does not contact the casing. When the box press reaches the home position, the wheels drop into the recesses which are disposed at the ends of the grooves, the valve plate is carried on the valve opening, and the box press presses the valve plate to increase the tightness. The valve plate does not suffer wear as it does not over other parts. (Yamaguchi, T.)

  5. Sequential transcatheter aortic valve implantation due to valve dislodgement

    DEFF Research Database (Denmark)

    Campante Teles, Rui; Costa, Cátia; Almeida, Manuel

    2017-01-01

    Transcatheter aortic valve implantation (TAVI) has become an important treatment in high surgical risk patients with severe aortic stenosis (AS), whose complications need to be managed promptly. The authors report the case of an 86-year-old woman presenting with severe symptomatic AS, rejected...

  6. A Calculation of hydrodynamic noise of control valve on instrumentation and control system using smart plant

    International Nuclear Information System (INIS)

    Demon Handoyo; Djoko H Nugroho

    2012-01-01

    It has been calculated characteristics of the control valve Instrumentation and Control Systems using Smart Plant software. This calculation is done in order to control the valve that will be installed as part of the instrumentation and control systems to provide the performance according to the design. The characteristics that have been calculated are Reynolds number factors which are related to the flow regime in the valve. Critical pressure factor, Valve Hydrodynamic cavitation and noise index. In this paper the discussion will be limited to matters relating to Hydrodynamic noise generation process using model of the instrumentation and control system in the plant design in yellow cake PIPKPP activities in 2012. The results of the calculation of the noise on the valves design are in the range between 9.58~70.1 dBA. (author)

  7. Supra-annular valve strategy for an early degenerated transcatheter balloon-expandable heart valve.

    Science.gov (United States)

    Kamioka, Norihiko; Caughron, Hope; Corrigan, Frank; Block, Peter; Babaliaros, Vasilis

    2018-01-23

    Currently, there are no recommendations regarding the selection of valve type for a transcatheter heart valve (THV)-in-THV procedure. A supra-annular valve design may be superior in that it results in a larger effective orifice area and may have a lower chance of valve thrombosis after THV-in-THV. In this report, we describe the use of a supra-annular valve strategy for an early degenerated THV. © 2018 Wiley Periodicals, Inc.

  8. Heart valve bioprosthesis durability: a challenge to the new generation of porcine valves.

    Science.gov (United States)

    Valente, M; Minarini, M; Maizza, A F; Bortolotti, U; Thiene, G

    1992-01-01

    Long-term experience with first generation porcine valve xenografts enabled identification of the major limitations to their durability: (1) prosthetic-ventricular mismatch due to the high profile of the stent in patients with mitral stenosis and a small left ventricle; (2) high-pressure fixation with loss of natural collagen crimping in the fibrosa, and wash-out of proteoglycans in the spongiosa; (3) xenograft tissue autolysis, due to the long interval between animal slaughter and aortic valve removal fixation; (4) muscle shelf in the right coronary cusp, which created a gradient and could undergo accelerated calcification and/or spontaneous perforation with time; (5) a flexible polypropylene stent, which could creep or even fracture with consequent inward bending of the stent; (6) progressive time-related dystrophic calcification; (7) host fibrous tissue ingrowth. An awareness of these limitations stimulated technical modifications, which frequently brought about distinct improvements: (1) the reduction of the stent profile eliminated the problem of mismatch, but resulted in a higher tendency towards cusp prolapse and earlier commissural tearing; (2) natural collagen waviness, proteoglycans and cusp extensibility were preserved by employing low or even zero pressure during the fixation process; (3) earlier valve fixation enabled preservation of cell integrity; (4) a new orifice for small valves was designed by replacing the right muscular cusp, thus achieving less gradient and avoiding muscle-shelf-related complications; (5) polypropylene was replaced by Delrin as stent material; (6) calcium-retarding agents like T6 and toluidine blue were applied during commercial processing and storage in order to mitigate tissue mineralization.(ABSTRACT TRUNCATED AT 250 WORDS)

  9. Acute Right Coronary Ostial Stenosis during Aortic Valve Replacement

    Directory of Open Access Journals (Sweden)

    Sarwar Umran

    2012-01-01

    Full Text Available We report a rare case of acute right coronary artery stenosis developing in a patient undergoing aortic valve replacement. We present a case report with a brief overview of the literature relating to coronary artery occlusion associated with cardiac valve surgery - the theories and treatments are discussed. A 85 year-old female was admitted under the care of the cardiothoracic team with signs and symptoms of heart failure. Investigations, including cardiac echocardiography and coronary angiography, indicated a critical aortic valve stenosis. Intraoperative right ventricular failure ensued post aortic valve replacement. Subsequent investigations revealed an acute occlusion of the proximal right coronary artery with resultant absence of distal flow supplying the right ventricle. An immediate right coronary artery bypass procedure was performed with resolution of the right ventricular failure. Subsequent weaning off cardiopulmonary bypass was uneventful and the patient continued to make excellent recovery in the postoperative phase. To our knowledge this is one of the few documented cases of intraoperative acute coronary artery occlusion developing during valve surgery. However, surgeons should be aware of the potential for acute occlusion so that early recognition and rapid intervention can be instituted.

  10. Outlet strut fracture of Björk-Shiley convexo-concave valves: can valve-manufacturing characteristics explain the risk?

    Science.gov (United States)

    Omar, R Z; Morton, L S; Beirne, M; Blot, W J; Lawford, P V; Hose, R; Taylor, K M

    2001-06-01

    Björk-Shiley 60 degrees convexo-concave prosthetic heart valves (Shiley, Inc, Irvine, Calif, a subsidiary of Pfizer, Inc) continue to be a concern for approximately 35,000 nonexplanted patients worldwide, with approximately 600 events reported to the manufacturer to date. Fractures of the outlet struts of the valves began to appear in the early 1980s and have continued to the present, but their causes are only partially understood. A matched case-control study was conducted evaluating manufacturing records for 52 valves with outlet strut fractures and 248 control subjects matched for age at implantation, valve size, and valve position. In addition to the risk factors recognized as determinants of outlet strut fracture, the United Kingdom case-control study has observed 7- to 9-fold increased risk with performance of multiple hook deflection tests. This test was performed more than once, usually after rework on the valve. Six valves in this study underwent multiple hook deflection tests, of which 4 experienced an outlet strut fracture. Cracks and further rework were noted for these valves. Significant associations were also observed between outlet strut fracture and disc-to-strut gap measurements taken before the attachment of the sewing ring. It is our view that a combination of factors related to valve design, manufacturing process, and patient characteristics are responsible for outlet strut fractures of Björk-Shiley convexo-concave valves. Multiple hook deflection tests have emerged as a potential new risk factor for outlet strut fracture in both The Netherlands and the United Kingdom. This factor appears to be correlated with the presence of other abnormalities. A further study is needed to investigate the factors correlated with multiple hook deflection tests. On confirmation of risk, the presence of multiple hook deflection tests may be added to equations, quantifying the risk of outlet strut fracture for comparison against risk of mortality and serious

  11. Aortic valve function after bicuspidization of the unicuspid aortic valve.

    Science.gov (United States)

    Aicher, Diana; Bewarder, Moritz; Kindermann, Michael; Abdul-Khalique, Hashim; Schäfers, Hans-Joachim

    2013-05-01

    Unicuspid aortic valve (UAV) anatomy leads to dysfunction of the valve in young individuals. We introduced a reconstructive technique of bicuspidizing the UAV. Initially we copied the typical asymmetry of a normal bicuspid aortic valve (BAV) (I), later we created a symmetric BAV (II). This study compared the hemodynamic function of the two designs of a bicuspidized UAV. Aortic valve function was studied at rest and during exercise in 28 patients after repair of UAV (group I, n = 8; group II, n = 20). There were no differences among the groups I and II with respect to gender, age, body size, or weight. All patients were in New York Heart Association class I. Six healthy adults served as control individuals. All patients were studied with transthoracic echocardiography between 4 and 65 months postoperatively. Systolic gradients were assessed by continuous wave Doppler while patients were at rest and exercising on a bicycle ergometer. Aortic regurgitation was grade I or less in all patients. Resting gradients were significantly elevated in group I compared with group II and control individuals (group I, peak 33.8 ± 7.8 mm Hg; mean 19.1 ± 5.4 mm Hg; group II, peak 15.8 ± 5.4, mean 8.2 ± 2.8 mm Hg; control individuals, peak 6.0 ± 1.6, mean 3.2 ± 0.8 mm Hg; p competence. A symmetric repair design leads to improved systolic aortic valve function at rest and during exercise. Copyright © 2013 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

  12. Pump and valve research at the Oak Ridge National Laboratory

    International Nuclear Information System (INIS)

    Haynes, H.D.

    1992-01-01

    Over the last several years, the Oak Ridge National Laboratory (ORNL) has carried out several aging assessments on pumps and valves under the NRC's Nuclear Plant Aging Research (NPAR) Program. In addition, ORNL has established an Advanced Diagnostic Engineering Research and Development Center (ADEC) in order to play a key role in the field of diagnostic engineering. Initial ADEC research projects have addressed problems that were identified, at least in part, by the NPAR and other NRC-sponsored programs. This paper summarizes the pump and valve related research that has been done at ORNL and describes in more detail several diagnostic techniques developed at ORNL which are new commercially available

  13. Optothermally actuated capillary burst valve

    DEFF Research Database (Denmark)

    Eriksen, Johan; Bilenberg, Brian; Kristensen, Anders

    2017-01-01

    be burst by raising the temperature due to the temperature dependence of the fluid surface tension. We address individual valves by using a local heating platform based on a thin film of near infrared absorber dye embedded in the lid used to seal the microfluidic device [L. H. Thamdrup et al., Nano Lett...

  14. Spring valve for well completion

    Energy Technology Data Exchange (ETDEWEB)

    Gorbatov, P T

    1966-07-22

    A spring-loaded valve for well completion consists of a housing with a spring-loaded closing element. In order to protect the closing element from corrosion which might lower the pressure drop, the closing element is made in the form of a piston. It is tightly connected with sealing elements. The housing has orifices, overlapping the piston in the initial position.

  15. Hydraulic servo control spool valve

    Science.gov (United States)

    Miller, Donald M.

    1983-01-01

    A servo operated spool valve having a fixed sleeve and axially movable spool. The sleeve is machined in two halves to form a long, narrow tapered orifice slot across which a transverse wall of the spool is positioned. The axial position of the spool wall along the slot regulates the open orifice area with extreme precision.

  16. Secreted Klotho Attenuates Inflammation-Associated Aortic Valve Fibrosis in Senescence-Accelerated Mice P1.

    Science.gov (United States)

    Chen, Jianglei; Fan, Jun; Wang, Shirley; Sun, Zhongjie

    2018-05-01

    Senescence-accelerated mice P1 (SAMP1) is an aging model characterized by shortened lifespan and early signs of senescence. Klotho is an aging-suppressor gene. The purpose of this study is to investigate whether in vivo expression of secreted klotho ( Skl ) gene attenuates aortic valve fibrosis in SAMP1 mice. SAMP1 mice and age-matched (AKR/J) control mice were used. SAMP1 mice developed obvious fibrosis in aortic valves, namely fibrotic aortic valve disease. Serum level of Skl was decreased drastically in SAMP1 mice. Expression of MCP-1 (monocyte chemoattractant protein 1), ICAM-1 (intercellular adhesion molecule 1), F4/80, and CD68 was increased in aortic valves of SAMP1 mice, indicating inflammation. An increase in expression of α-smooth muscle actin (myofibroblast marker), transforming growth factorβ-1, and scleraxis (a transcription factor of collagen synthesis) was also found in aortic valves of SAMP1 mice, suggesting that accelerated aging is associated with myofibroblast transition and collagen gene activation. We constructed adeno-associated virus 2 carrying mouse Skl cDNA for in vivo expression of Skl. Skl gene delivery effectively increased serum Skl of SAMP1 mice to the control level. Skl gene delivery inhibited inflammation and myofibroblastic transition in aortic valves and attenuated fibrotic aortic valve disease in SAMP1 mice. It is concluded that senescence-related fibrotic aortic valve disease in SAMP1 mice is associated with a decrease in serum klotho leading to inflammation, including macrophage infiltration and transforming growth factorβ-1/scleraxis-driven myofibroblast differentiation in aortic valves. Restoration of serum Skl levels by adeno-associated virus 2 carrying mouse Skl cDNA effectively suppresses inflammation and myofibroblastic transition and attenuates aortic valve fibrosis. Skl may be a potential therapeutic target for fibrotic aortic valve disease. © 2018 American Heart Association, Inc.

  17. Effects on fatigue life of gate valves due to higher torque switch settings during operability testing

    International Nuclear Information System (INIS)

    Richins, W.D.; Snow, S.D.; Miller, G.K.; Russell, M.J.; Ware, A.G.

    1995-12-01

    Some motor operated valves now have higher torque switch settings due to regulatory requirements to ensure valve operability with appropriate margins at design basis conditions. Verifying operability with these settings imposes higher stem loads during periodic inservice testing. These higher test loads increase stresses in the various valve internal parts which may in turn increase the fatigue usage factors. This increased fatigue is judged to be a concern primarily in the valve disks, seats, yokes, stems, and stem nuts. Although the motor operators may also have significantly increased loading, they are being evaluated by the manufacturers and are beyond the scope of this study. Two gate valves representative of both relatively weak and strong valves commonly used in commercial nuclear applications were selected for fatigue analyses. Detailed dimensional and test data were available for both valves from previous studies at the Idaho National Engineering Laboratory. Finite element models were developed to estimate maximum stresses in the internal parts of the valves and to identity the critical areas within the valves where fatigue may be a concern. Loads were estimated using industry standard equations for calculating torque switch settings prior and subsequent to the testing requirements of USNRC Generic Letter 89--10. Test data were used to determine both; (1) the overshoot load between torque switch trip and final seating of the disk during valve closing and (2) the stem thrust required to open the valves. The ranges of peak stresses thus determined were then used to estimate the increase in the fatigue usage factors due to the higher stem thrust loads. The usages that would be accumulated by 100 base cycles plus one or eight test cycles per year over 40 and 60 years of operation were calculated

  18. Valve assembly having remotely replaceable bearings

    International Nuclear Information System (INIS)

    Johnson, E.R.; Tanner, D.E.

    1980-01-01

    A valve assembly having remotely replaceable bearings is disclosed wherein a valve disc is supported within a flow duct for rotation about a pair of axially aligned bearings, one of which is carried by a spindle received within a diametral bore in the valve disc, and the other of which is carried by a bearing support block releasably mounted on the duct circumferentially of an annular collar on the valve disc coaxial with its diametrical bore. The spindle and bearing support block are adapted for remote removal to facilitate servicing or replacement of the valve disc support bearings

  19. Multifunctional four-port directional control valve constructed from logic valves

    International Nuclear Information System (INIS)

    Lisowski, E.; Czyżycki, W.; Rajda, J.

    2014-01-01

    Highlights: • Directional valve with standard ISO 440-08 has been constructed from logic valves. • Only one innovative valve may replace whole family of the standard valves. • CFD analysis and bench tests of the innovative valve has been carried. • Parameters of the innovative valve are equaling or surpassing the standard ones. • The innovative valve has additional possibilities of pressure and flow control. - Abstract: The paper refers to four-port solenoid pilot operated valves, which are subplate mounted in a hydraulic system in accordance with the ISO 4401 standard. Their widespread use in many machines and devices causes a continuing interest in the development of their design by both the scientific centers and the industry. This paper presents an innovative directional control valve based on the use of logic valves and a methodology followed for the design of it by using Solid Edge CAD and ANSYS/Fluent CFD software. The valve design methodology takes into account the need to seek solutions that minimize flow resistance through the valve. For this purpose, the flow paths are prepared by means of CAD software and pressure-flow curves are determined as a result of CFD analysis. The obtained curves are compared with the curves available in the catalogs of spool type directional control valves. The new solution allows to replace the whole family of spool type four-port directional control valves by one valve built of logic valves. In addition, the innovative directional control valve provides leak-proof shutting the flow paths off and also it can control flow rate and even pressure of working liquid. A prototype of the valve designed by the presented method has been made and tested on the test bench. The results quoted in the paper confirm that the developed logic type directional control valve is able to meet all designed connection configurations, and the obtained pressure-flow curves show very good conformity with the results of CFD analysis

  20. Simple versus complex degenerative mitral valve disease.

    Science.gov (United States)

    Javadikasgari, Hoda; Mihaljevic, Tomislav; Suri, Rakesh M; Svensson, Lars G; Navia, Jose L; Wang, Robert Z; Tappuni, Bassman; Lowry, Ashley M; McCurry, Kenneth R; Blackstone, Eugene H; Desai, Milind Y; Mick, Stephanie L; Gillinov, A Marc

    2018-07-01

    At a center where surgeons favor mitral valve (MV) repair for all subsets of leaflet prolapse, we compared results of patients undergoing repair for simple versus complex degenerative MV disease. From January 1985 to January 2016, 6153 patients underwent primary isolated MV repair for degenerative disease, 3101 patients underwent primary isolated MV repair for simple disease (posterior prolapse), and 3052 patients underwent primary isolated MV repair for complex disease (anterior or bileaflet prolapse), based on preoperative echocardiographic images. Logistic regression analysis was used to generate propensity scores for risk-adjusted comparisons (n = 2065 matched pairs). Durability was assessed by longitudinal recurrence of mitral regurgitation and reoperation. Compared with patients with simple disease, those undergoing repair of complex pathology were more likely to be younger and female (both P values < .0001) but with similar symptoms (P = .3). The most common repair technique was ring/band annuloplasty (3055/99% simple vs 3000/98% complex; P = .5), followed by leaflet resection (2802/90% simple vs 2249/74% complex; P < .0001). Among propensity-matched patients, recurrence of severe mitral regurgitation 10 years after repair was 6.2% for simple pathology versus 11% for complex pathology (P = .007), reoperation at 18 years was 6.3% for simple pathology versus 11% for complex pathology, and 20-year survival was 62% for simple pathology versus 61% for complex pathology (P = .6). Early surgical intervention has become more common in patients with degenerative MV disease, regardless of valve prolapse complexity or symptom status. Valve repair was associated with similarly low operative risk and time-related survival but less durability in complex disease. Lifelong annual echocardiographic surveillance after MV repair is recommended, particularly in patients with complex disease. Copyright © 2018 The American Association for Thoracic Surgery

  1. Valve for closing a steam line

    International Nuclear Information System (INIS)

    Meyer, W.; Potrykus, G.

    1976-01-01

    Instead of several control elements, the quick-closing valve, especially in the main-steam line between steam generator and turbine of a power station has the valve cone itself as the only movable part, acting with its inner surface as a piston within a second cylinder space. The valve shaft is at the same time a piston rod with a stepped piston at the upper end. This piston is loaded in a cylinder at the upspace below the valve cover on one hand by a spring, on the other hand by its own medium. Two non-return valves, one of it in a bore of the valve cone, connect the first-mentioned cylinder space with the steam-loaded inlet resp. outlet side of the valve. For controlling the valve, a magnet valve is sufficient. By automatic control of the valve cone coupled with several pistons several control lines can be omitted. There are also no pressurized control lines outside the valve which could be damaged by exterior influences. (ERA) [de

  2. Mechanical versus bioprosthetic aortic valve replacement.

    Science.gov (United States)

    Head, Stuart J; Çelik, Mevlüt; Kappetein, A Pieter

    2017-07-21

    Mechanical valves used for aortic valve replacement (AVR) continue to be associated with bleeding risks because of anticoagulation therapy, while bioprosthetic valves are at risk of structural valve deterioration requiring reoperation. This risk/benefit ratio of mechanical and bioprosthetic valves has led American and European guidelines on valvular heart disease to be consistent in recommending the use of mechanical prostheses in patients younger than 60 years of age. Despite these recommendations, the use of bioprosthetic valves has significantly increased over the last decades in all age groups. A systematic review of manuscripts applying propensity-matching or multivariable analysis to compare the usage of mechanical vs. bioprosthetic valves found either similar outcomes between the two types of valves or favourable outcomes with mechanical prostheses, particularly in younger patients. The risk/benefit ratio and choice of valves will be impacted by developments in valve designs, anticoagulation therapy, reducing the required international normalized ratio, and transcatheter and minimally invasive procedures. However, there is currently no evidence to support lowering the age threshold for implanting a bioprosthesis. Physicians in the Heart Team and patients should be cautious in pursuing more bioprosthetic valve use until its benefit is clearly proven in middle-aged patients. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2017. For permissions, please email: journals.permissions@oup.com.

  3. Impact of patient-prosthesis mismatch after transcatheter aortic valve-in-valve implantation in degenerated bioprostheses.

    Science.gov (United States)

    Seiffert, Moritz; Conradi, Lenard; Baldus, Stephan; Knap, Malgorzata; Schirmer, Johannes; Franzen, Olaf; Koschyk, Dietmar; Meinertz, Thomas; Reichenspurner, Hermann; Treede, Hendrik

    2012-03-01

    Transcatheter valve-in-valve implantation is evolving as an alternative to reoperative valve replacement in high-risk patients with degenerated bioprostheses. Nevertheless, hemodynamic performance is limited by the previously implanted xenograft. We report our experience with patient-prosthesis mismatch (PPM) after valve-in-valve implantation in the aortic position. Eleven patients (aged 79.3 ± 6.1 years) received transapical implantation of a balloon-expandable pericardial heart valve into a degenerated bioprosthesis (size, 23.9 ± 1.6 mm; range, 21-27 mm) in the aortic position. All patients were considered high risk for surgical valve replacement (logistic European System for Cardiac Operative Risk Evaluation, 31.8% ± 24.1%). Severe PPM was defined as an indexed effective orifice area less than 0.65 cm(2)/m(2), determined by discharge echocardiography. Severe PPM was evident in 5 patients (group 1) and absent in 6 patients (group 2). Mean transvalvular gradients decreased from 29.2 ± 15.4 mm Hg before implantation to 21.2 ± 9.7 mm Hg at discharge (group 1) and from 28.2 ± 9.0 mm Hg before implantation to 15.2 ± 6.5 mm Hg at discharge (group 2). Indexed effective orifice area increased from 0.5 ± 0.1 cm(2)/m(2) to 0.6 ± 0.1 cm(2)/m(2) and from 0.6 ± 0.3 cm(2)/m(2) to 0.8 ± 0.3 cm(2)/m(2). Aortic regurgitation decreased from grade 2.0 ± 1.1 to 0.4 ± 0.5 overall. No differences in New York Heart Association class improvement or survival during follow-up were observed. One patient required reoperation for symptomatic PPM 426 days after implantation. Valve-in-valve implantation can be performed in high-risk surgical patients to avoid reoperation. However, PPM frequently occurs, making adequate patient selection crucial. Small bioprostheses (body surface area less than 1.8 m(2). Larger prostheses seem to carry a lower risk for PPM. Although no delay in clinical improvement was seen at short-term, 1 PPM-related surgical intervention raises concern regarding

  4. Acoustic valve leak detection in nuclear plants

    International Nuclear Information System (INIS)

    Dimmick, J.G.; Dickey, J.W.

    1983-01-01

    Internal valve leakage is a hidden energy loss and can cause or prolong a forced outage. Recent advances in acoustic detection of internal valve leakage have reduced piping system maintenance costs, unnecessary downtime, and energy waste. Extremely short payback periods have been reported by plants applying this technology to preventive maintenance, troubleshooting, energy conservation and outage planning. Sensors temporarily attached to the outside of valves and connected to the instruments detect ultrasonic acoustic emissions which are characteristic of internal valve leakage. Since the sensors are attached to the outside of the valves, the time and expense of dismantling the valves or removing them from the systems are eliminated. This paper describes the instrumentation and specific applications to nuclear plant valves, including independent verification of initial findings. Guidelines for potential users, including instrumentation selection, training requirements, application planning, and the choice of in-house versus contract services are discussed

  5. The effect of varying degrees of stenosis on the characteristics of turbulent pulsatile flow through heart valves.

    Science.gov (United States)

    Bluestein, D; Einav, S

    1995-08-01

    Many problems and complications associated with heart valves are related to the dynamic behavior of the valve and the resultant unsteady flow patterns. An accurate depiction of the spatial and temporal velocity and rms distributions imparts better understanding of flow related valve complications, and may be used as a guideline in valve design. While the generalized correlation between increased turbulence level and the severity of the stenosis is well established, few studies addressed the issue of the intermittent nature of turbulence and its timing in the cardiac cycle, and almost none assessed the effect of a progressive stenosis on the flow characteristics through heart valves. In this experimental work we simulated the type of flow which is present in normal and stenosed valves and conducted a comprehensive investigation of valve hemodynamics, valvular turbulence and morphology under varying degrees of stenosis. The characteristics of valves and stenoses were simulated closely, to achieve the flow conditions that initiate turbulent flow conditions. Laser Doppler anemometry (LDA) measurements were carried out in a pulse duplicator system distal to trileaflet polyurethane prosthetic heart valves, installed at mitral and aortic positions. The effect of the degree of the stenosis was comparatively studied through the structure of the turbulent jets emerging from normal and stenotic heart valves. Maximum turbulence level was achieved during the decelerating phase and correlated to the severity of the stenosis, followed by relaminarization of the flow during the acceleration phase. The intermittent nature of the turbulence emphasized the importance of realizing the timing of the turbulence production and its spatial location for optimizing current valve designs. The plug flow through the normal aortic valve prosthesis was replaced by jet like behavior for a 65% stenosis, with the jet becoming narrower and stronger for a 90% stenosis. The morphology of the velocity

  6. An improved gate valve for critical applications in nuclear power plants

    Energy Technology Data Exchange (ETDEWEB)

    Kalsi, M.S.; Alvarez, P.D.; Wang, J.K.; Somagyi, D. [Kalsi Engineering, Inc., Sugar Land, TX (United States)] [and others

    1996-12-01

    U.S. Nuclear Regulatory Commission Generic Letters 89-10 for motor-operated valves (MOVs) and 95-07 for all power-operated valves document in detail the problems related to the performance of the safety-related valves in nuclear power plants. The problems relate to lack of reliable operation under design basis conditions including higher than anticipated stem thrust, unpredictable valve behavior, damage to the valve internals under blowdown/high flow conditions, significant degradation of performance when cycled under AP and flow, thermal binding, and pressure locking. This paper describes an improved motor-operated flexible wedge gate valve design, the GE Sentinel Valve, which is the outcome of a comprehensive and systematic development effort undertaken to resolve the issues identified in the NRC Generic Letters 89-10 and 95-07. The new design provides a reliable, long-term, low maintenance cost solution to the nuclear power industry. One of the key features incorporated in the disc permits the disc flexibility to be varied independently of the disc thickness (pressure boundary) dictated by the ASME Section III Pressure Vessel & Piping Code stress criteria. This feature allows the desired flexibility to be incorporated in the disc, thus eliminating thermal binding problems. A matrix of analyses was performed using finite element and computational fluid dynamics approaches to optimize design for stresses, flexibility, leak-tightness, fluid flow, and thermal effects. The design of the entire product line was based upon a consistent set of analyses and design rules which permit scaling to different valve sizes and pressure classes within the product line. The valve meets all of the ASME Section III Code design criteria and the N-Stamp requirements. The performance of the valve was validated by performing extensive separate effects and plant in-situ tests. This paper summarizes the key design features, analyses, and test results.

  7. An improved gate valve for critical applications in nuclear power plants

    International Nuclear Information System (INIS)

    Kalsi, M.S.; Alvarez, P.D.; Wang, J.K.; Somagyi, D.

    1996-01-01

    U.S. Nuclear Regulatory Commission Generic Letters 89-10 for motor-operated valves (MOVs) and 95-07 for all power-operated valves document in detail the problems related to the performance of the safety-related valves in nuclear power plants. The problems relate to lack of reliable operation under design basis conditions including higher than anticipated stem thrust, unpredictable valve behavior, damage to the valve internals under blowdown/high flow conditions, significant degradation of performance when cycled under AP and flow, thermal binding, and pressure locking. This paper describes an improved motor-operated flexible wedge gate valve design, the GE Sentinel Valve, which is the outcome of a comprehensive and systematic development effort undertaken to resolve the issues identified in the NRC Generic Letters 89-10 and 95-07. The new design provides a reliable, long-term, low maintenance cost solution to the nuclear power industry. One of the key features incorporated in the disc permits the disc flexibility to be varied independently of the disc thickness (pressure boundary) dictated by the ASME Section III Pressure Vessel ampersand Piping Code stress criteria. This feature allows the desired flexibility to be incorporated in the disc, thus eliminating thermal binding problems. A matrix of analyses was performed using finite element and computational fluid dynamics approaches to optimize design for stresses, flexibility, leak-tightness, fluid flow, and thermal effects. The design of the entire product line was based upon a consistent set of analyses and design rules which permit scaling to different valve sizes and pressure classes within the product line. The valve meets all of the ASME Section III Code design criteria and the N-Stamp requirements. The performance of the valve was validated by performing extensive separate effects and plant in-situ tests. This paper summarizes the key design features, analyses, and test results

  8. Thromboembolism and mechanical heart valves : A randomized study revisited

    NARCIS (Netherlands)

    Kuntze, CEE; Blackstone, EH; Ebels, T

    Background. This study was designed to revise and substantiate previous inferences, based on short-term follow-up, about differences in the incidence of anticoagulant-related events after heart valve replacement among patients who had been randomly assigned to receive either a Bjork-Shiley,

  9. Valve testing for UK PWR safety applications

    International Nuclear Information System (INIS)

    George, P.T.; Bryant, S.

    1989-01-01

    Extensive testing and development has been done by the Central Electricity Generating Board (CEGB) to support the design, construction and operation of Sizewell B, the UK's first PWR. A Blowdown Rig for the Assessment of Valve Operability - (BRAVO) has been constructed at the CEGB Marchwood Engineering Laboratory to reproduce PWR Pressurizer fluid conditions for the full scale testing of Pressurizer Relief System (PRS) valves. A full size tandem pair of Pilot Operated Safety Relief Valves (POSRVs) is being tested under the full range of pressurizer fluid conditions. Tests to date have produced important data on the performance of the valve in its Cold Overpressure protection mode of operation and on methods for the in-service testing of the valve. Also, a full size pressurizer safety valve has been tested under full PRS fluid conditions to develop a methodology for the pre-service testing of the Sizewell valves. Further work will be carried out to develop procedures for the in-service testing of the valve. In the Main Steam Safety Valve test program carried out at the Siemens-KWU Test Facilities, a single MSSV from three potential suppliers was tested under full secondary system conditions. The test results have been analyzed and are reflected in the CEGB's arrangements for the pre-service and in-service testing of the Sizewell MSSVs. Valves required to interrupt pipebreak flow must be qualified for this duty by testing or a combination of testing and analysis. To obtain guidance on the performance of such tests gate and globe valves have been subjected to simulated pipebreaks under PWR primary circuit conditions. In the light of problems encountered with gate valve closure under these conditions, further tests are currently being carried out on the BRAVO facility on a gate valve, in preparation for the full scale flow interruption qualification testing of the Sizewell main steam isolation valve

  10. Analysis on typical illegal events for nuclear safety class 1 valve

    International Nuclear Information System (INIS)

    Tian Dongqing; Gao Runsheng; Jiao Dianhui; Yang Lili; Chen Peng

    2014-01-01

    Illegal welding events of nuclear safety class l valve forging occurred to the manufacturer, while the valve was returned to be repaired. Illegal nondestructive test event of nuclear safety class valve occurred also to the manufacturer in the manufacturing process. The two events have resulted in quality incipient fault for the installed valves and the valves in the manufacturing process. It was reflected that operation of the factory quality assurance system isn't activated, and nuclear power engineering and operating company have insufficient supervision. The event-related parties should strengthen quality management and process control, get rid of the quality incipient fault, and experience feedback should be done well to guarantee quality of equipment in nuclear power plant. (authors)

  11. Design Method for Fast Switching Seat Valves for Digital Displacement Machines

    DEFF Research Database (Denmark)

    Roemer, Daniel Beck; Johansen, Per; Pedersen, Henrik C.

    2014-01-01

    corresponding to the piston movement, which has been shown to facilitate superior part load efficiency combined with high bandwidth compared to traditional displacement machines. However, DD machines need fast switching on-off valves with low pressure loss for efficient operation, especially in fast rotating......Digital Displacement (DD) machines are upcoming technology where the displacement of each pressure chamber is controlled electronically by use of two fast switching seat valves. The effective displacement and operation type (pumping/motoring) may be controlled by manipulating the seat valves...... method for DD seat valves are presented, taking into account the significant aspects related to obtaining efficient DD valves with basis in a given DD machine specifications. The seat area is minimized and the stroke length is minimized to obtain fast switching times while considering the pressure loss...

  12. Transcatheter aortic valve prosthesis surgically replaced 4 months after implantation

    DEFF Research Database (Denmark)

    Thyregod, Hans Gustav; Lund, Jens Teglgaard; Engstrøm, Thomas

    2010-01-01

    Transcatheter aortic valve implantation is a new and rapidly evolving treatment option for high-risk surgical patients with degenerative aortic valve stenosis. Long-term results with these new valve prostheses are lacking, and potential valve dysfunction and failure would require valve replacemen....... We report the first case of surgical valve replacement in a patient with a dysfunctional transcatheter-implanted aortic valve prosthesis 4 months after implantation....

  13. Testing of cobalt-free alloys for valve applications using a special test loop

    International Nuclear Information System (INIS)

    Benhamou, C.

    1992-01-01

    Considering that use of cobalt alloys should be avoided as far as possible in PWR components, a programme aimed at establishing the performance of cobalt-free alloys has been performed for valve applications, where cobalt alloys are mainly used. Referring to past work, two types of cobalt-free alloys were selected: Ni-Cr-B-Si and Ni-Cr-Fe alloys. Cobalt-free valves' behaviour has been evaluated comparatively with cobalt valves by implementation of a programme in a special PWR test loop. At the issue of the loop test programme, which included endurance, thermal shock and erosion tests, cobalt-free alloys candidate to replace cobalt alloys are proposed in relation with valve type (globe valve and swing check valve). The following was established: (i) Colmonoy 4-26 (Ni-Cr-B-Si alloy) and Cenium Z20 (Ni-Cr-Fe alloy) deposited by plasma arc process were found suitable for use in 3inch swing check valves; (ii) for integral parts acting as guide rings, Nitronic 60 and Cesium Z20/698 were tested successfully; (iii) for small-bore components such as 2inch globe valves, no solution can yet be proposed; introduction of cobalt-free alloys is dependent on the development of automatic advanced arc surfacing techniques applied to small-bore components

  14. Analysis of design parameters and flow characteristics of the vortex valve for SIT flow control

    International Nuclear Information System (INIS)

    Hwang, Young Dong; Chang, Moon Hee; Kim, Seong O.; Kim, Young In.

    1997-01-01

    This study was performed to provide a technical basis for the development of the vortex valve which will be adopted in Korean Advanced Reactor. The influence of nondimensional and geometrical parameters of the vortex valve were investigated by analyzing the flow field of the vortex chamber, and the performance related parameters were evaluated by utilizing of the published experimental and analytical data. Also the level transients of the stand pipe were investigated by using of the simplified analytical model. In order to obtain the more detailed information on the vortex flow field, three dimensional preliminary analyses for the vortex valve design were conducted by FLUENT code. This study were carried out by using the simplified analytical model of the vortex valve and downstream pipe. However, the detailed analysis on the integrated system of the vortex valve with the as built design data and the required operating conditions should be performed to obtain the more accurate results on the vortex valve behavior. Also the experimental study over a wide range of operating conditions to develop the correlation of the design parameters and the performance verification should be performed for the practical design and engineering applications of the vortex valve. The results of this study can be used as a basic information for the development of the vortex valve design for the SIT of Korean Advanced Reactor. (author). 12 refs., 5 tabs., 33 figs

  15. Experimental study on the effect of an artificial cardiac valve on the left ventricular flow

    Science.gov (United States)

    Wang, JiangSheng; Gao, Qi; Wei, RunJie; Wang, JinJun

    2017-09-01

    The use of artificial valves to replace diseased human heart valves is currently the main solution to address the malfunctioning of these valves. However, the effect of artificial valves on the ventricular flow still needs to be understood in flow physics. The left ventricular flow downstream of a St. Jude Medical (SJM) bileaflet mechanical heart valve (BMHV), which is a widely implanted mechanical bileaflet valve, is investigated with time-resolved particle image velocimetry in the current work. A tilting-disk valve is installed on the aortic orifice to guarantee unidirectional flow. Several post-processing tools are applied to provide combined analyses of the physics involved in the ventricular flow. The triple jet pattern that is closely related to the characteristics of the bileaflet valve is discussed in detail from both Eulerian and Lagrangian views. The effects of large-scale vortices on the transportation of blood are revealed by the combined analysis of the tracking of Lagrangian coherent structures, the Eulerian monitoring of the shear stresses, and virtual dye visualization. It is found that the utilization of the SJM BMHV complicates the ventricular flow and could reduce the efficiency of blood transportation. In addition, the kinematics of the bileaflets is presented to explore the effects of flow structures on their motion. These combined analyses could elucidate the properties of SJM BMHV. Furthermore, they could provide new insights into the understanding of other complex blood flows.

  16. Comparison of Algorithms for the Optimal Location of Control Valves for Leakage Reduction in WDNs

    Directory of Open Access Journals (Sweden)

    Enrico Creaco

    2018-04-01

    Full Text Available The paper presents the comparison of two different algorithms for the optimal location of control valves for leakage reduction in water distribution networks (WDNs. The former is based on the sequential addition (SA of control valves. At the generic step Nval of SA, the search for the optimal combination of Nval valves is carried out, while containing the optimal combination of Nval − 1 valves found at the previous step. Therefore, only one new valve location is searched for at each step of SA, among all the remaining available locations. The latter algorithm consists of a multi-objective genetic algorithm (GA, in which valve locations are encoded inside individual genes. For the sake of consistency, the same embedded algorithm, based on iterated linear programming (LP, was used inside SA and GA, to search for the optimal valve settings at various time slots in the day. The results of applications to two WDNs show that SA and GA yield identical results for small values of Nval. When this number grows, the limitations of SA, related to its reduced exploration of the research space, emerge. In fact, for higher values of Nval, SA tends to produce less beneficial valve locations in terms of leakage abatement. However, the smaller computation time of SA may make this algorithm preferable in the case of large WDNs, for which the application of GA would be overly burdensome.

  17. Outcomes of Solo Smart valve in a single-center experience of 270 patients.

    Science.gov (United States)

    Liu, Hao; Khani-Hanjani, Abbas; Yang, Siyuan; Wang, Wei; Sidhu, Surita; Mullen, John; Modry, Dennis; Wang, Shaohua

    2018-04-03

    The Solo Smart pericardial aortic valve has been widely used in Europe as an option for aortic valve replacement (AVR). We are reporting early and midterm clinical outcomes of AVR with the Solo Smart valve in a single North America center. This is a retrospective study of 270 consecutive patients who had AVR at Mazankowski Alberta Heart Institute from February 2011 to March 2015. Follow-up and echocardiographic data were collected retrospectively from electronic and paper charts. Univariate and multivariate analysis were performed to evaluate the results. The mean age was 71.2±10.0 years, 67.4% were male, and 79.3% had combined procedures. Mean STS Score was 4.18±3.91. Early mortality was 3.7% for the entire group and 0% for isolated aortic valve replacement group. Mean cross-clamp time for isolated AVR and AVR with concomitant procedure was 70.8±12.7min and 117.0±45.0min, respectively. Permanent pacemaker implantation was necessary in 2.2% of patients. Echocardiography demonstrated a reduction in mean gradients from 40.8±17.4mmHg to 7.6±3.7 mmHg and peak gradient from 72.5 ± 48.8 mmHg to 15.5±7.5 mmHg. The 1-, 3-, and 5- year overall survival was 93.0%, 86.5% and 75.9%, respectively. At 5 years, freedom from valve-related death was 92.4%, freedom from structural valve deterioration and freedom from aortic valve reoperation were 96.4% and 98%, respectively. The Solo Smart valve is safe and has excellent hemodynamic performance. Aortic valve reoperation and rates of valve-related adverse events during midterm follow-up were low.

  18. Energetic and dynamic impact of counterbalance valves in fluid power machines

    International Nuclear Information System (INIS)

    Ritelli, Guido Francesco; Vacca, Andrea

    2013-01-01

    Highlights: • A graphical method to study the operation of counterbalance valves is presented. • The method is suitable to study the impact of valve settings on energy consumption. • Energetic considerations on a hydraulic crane were performed through numerical simulations. • Test were performed to validate the model on steady-state and transient conditions. • Energy consumption reduction potentials in current applications were discussed. - Abstract: Counterbalance valves negatively affect the dynamical behavior of most of the hydraulic systems in which they are utilized, and they introduce additional energy consumption. Various valve architectures have been developed to address the dynamical issues related to the use of these components, and empirical guidelines are also available to determine the best value of their set parameters such as the pilot ratio and the setting pressure. However, the problem of characterizing the portion of energy consumption introduced by counterbalance valves in a hydraulic circuit, and how this consumption is related to the valve setting, has never been addressed according to a systematic approach. This paper addresses this problem and proposes a general approach to study the energy consumption introduced by the counterbalance valves for given operating cycles of the machine in which they are used. In particular, a graphical method is utilized to study the effects of counterbalance valves in the whole hydraulic system. This method is also used to interpret the potentials for energy recovery associated with different settings of the counterbalance valves. The basic considerations which can be obtained using a graphical method are then supported by results provided by a more detailed numerical model of a particular reference machine, a hydraulic crane. The numerical model is validated on the basis of experimental results and permits to derive not only considerations about the energy consumption of counterbalance valves, but

  19. Depth of valve implantation, conduction disturbances and pacemaker implantation with CoreValve and CoreValve Accutrak system for Transcatheter Aortic Valve Implantation, a multi-center study.

    Science.gov (United States)

    Lenders, Guy D; Collas, Valérie; Hernandez, José Maria; Legrand, Victor; Danenberg, Haim D; den Heijer, Peter; Rodrigus, Inez E; Paelinck, Bernard P; Vrints, Christiaan J; Bosmans, Johan M

    2014-10-20

    Transcatheter Aortic Valve Implantation (TAVI) is now considered an indispensable treatment strategy in high operative risk patients with severe, symptomatic aortic stenosis. However, conduction disturbances and the need for Permanent Pacemaker (PPM) implantation after TAVI with the CoreValve prosthesis still remain frequent. We aimed to evaluate the implantation depth, the incidence and predictors of new conduction disturbances, and the need for PPM implantation within the first month after TAVI, using the new Accutrak CoreValve delivery system (ACV), compared to the previous generation CoreValve (non-ACV). In 5 experienced TAVI-centers, a total of 120 consecutive non-ACV and 112 consecutive ACV patients were included (n=232). The mean depth of valve implantation (DVI) was 8.4±4.0 mm in the non-ACV group and 7.1±4.0 mm in the ACV group (p=0.034). The combined incidence of new PPM implantation and new LBBB was 71.2% in the non-ACV group compared to 50.5% in the ACV group (p=0.014). DVI (p=0.002), first degree AV block (p=0.018) and RBBB (p<0.001) were independent predictors of PPM implantation. DVI (p<0.001) and pre-existing first degree AV-block (p=0.021) were identified as significant predictors of new LBBB. DVI is an independent predictor of TAVI-related conduction disturbances and can be reduced by using the newer CoreValve Accutrak delivery system, resulting in a significantly lower incidence of new LBBB and new PPM implantation. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  20. Application of ceramics to the sliding seat of valve bridge; Valve bridge yodobu eno ceramics tekiyo

    Energy Technology Data Exchange (ETDEWEB)

    Matsui, T; Ono, T [Mitsubishi Motors Corp., Tokyo (Japan)

    1997-10-01

    For use in the valve train, using an OHV (over head valve) configuration. of a 4 valve diesel engine for trucks and buses; we developed a valve bridge, a component of a valve train, with a ceramic head that is made of silicon nitride(Si3N4) in contact with a rocker arm in order to reduce cost and improve wear resistance for further diesel engine emissions regulations. In order to evaluate the effect of this valve bridge, RIG tests and durability tests on actual engines were carried out. 7 figs., 2 tabs.

  1. Performance of balanced bellows safety relief valves

    International Nuclear Information System (INIS)

    Lai, Y.S.

    1992-01-01

    By the nature of its design, the set point and lift of a conventional spring loaded safety relief valve are sensitive to back pressure. One way to reduce the adverse effects of the back pressure on the safety relief valve function is to install a balanced bellows in a safety relief valve. The metallic bellows has a rather wide range of manufacturing tolerance which makes the design of the bellows safety relief valve very complicated. The state-of-the-art balanced bellows safety relief valve can only substantially minimize, but cannot totally eliminate the back pressure effects on its set point and relieving capacity. Set point change is a linear function of the back pressure to the set pressure ratio. Depending on the valve design, the set point correction factor can be either greater or smaller than unity. There exists an allowable back pressure and critical back pressure for each safety relief valve. When total back pressure exceeds the R a , the relieving capacity will be reduced mainly resulting from the valve lift being reduced by the back pressure and the capacity reduction factor should be applied in valve sizing. Once the R c is exceeded, the safety relief valve becomes unstable and loses its over pressure protection capability. The capacity reduction factor is a function of system overpressure, but their relationship is non-linear in nature. (orig.)

  2. Traumatic Mitral Valve and Pericardial Injury

    Directory of Open Access Journals (Sweden)

    Nissar Shaikh

    2013-01-01

    Full Text Available Cardiac injury after blunt trauma is common but underreported. Common cardiac trauma after the blunt chest injury (BCI is cardiac contusion; it is very rare to have cardiac valve injury. The mitral valve injury during chest trauma occurs when extreme pressure is applied at early systole during the isovolumic contraction between the closure of the mitral valve and the opening of the aortic valve. Traumatic mitral valve injury can involve valve leaflet, chordae tendineae, or papillary muscles. For the diagnosis of mitral valve injury, a high index of suspicion is required, as in polytrauma patients, other obvious severe injuries will divert the attention of the treating physician. Clinical picture of patients with mitral valve injury may vary from none to cardiogenic shock. The echocardiogram is the main diagnostic modality of mitral valve injuries. Patient’s clinical condition will dictate the timing and type of surgery or medical therapy. We report a case of mitral valve and pericardial injury in a polytrauma patient, successfully treated in our intensive care unit.

  3. Simple Check Valves for Microfluidic Devices

    Science.gov (United States)

    Willis, Peter A.; Greer, Harold F.; Smith, J. Anthony

    2010-01-01

    A simple design concept for check valves has been adopted for microfluidic devices that consist mostly of (1) deformable fluorocarbon polymer membranes sandwiched between (2) borosilicate float glass wafers into which channels, valve seats, and holes have been etched. The first microfluidic devices in which these check valves are intended to be used are micro-capillary electrophoresis (microCE) devices undergoing development for use on Mars in detecting compounds indicative of life. In this application, it will be necessary to store some liquid samples in reservoirs in the devices for subsequent laboratory analysis, and check valves are needed to prevent cross-contamination of the samples. The simple check-valve design concept is also applicable to other microfluidic devices and to fluidic devices in general. These check valves are simplified microscopic versions of conventional rubber- flap check valves that are parts of numerous industrial and consumer products. These check valves are fabricated, not as separate components, but as integral parts of microfluidic devices. A check valve according to this concept consists of suitably shaped portions of a deformable membrane and the two glass wafers between which the membrane is sandwiched (see figure). The valve flap is formed by making an approximately semicircular cut in the membrane. The flap is centered over a hole in the lower glass wafer, through which hole the liquid in question is intended to flow upward into a wider hole, channel, or reservoir in the upper glass wafer. The radius of the cut exceeds the radius of the hole by an amount large enough to prevent settling of the flap into the hole. As in a conventional rubber-flap check valve, back pressure in the liquid pushes the flap against the valve seat (in this case, the valve seat is the adjacent surface of the lower glass wafer), thereby forming a seal that prevents backflow.

  4. Quantitative assessment of an aortic and pulmonary valve function according to valve fenestration

    International Nuclear Information System (INIS)

    Mirkhani, S.H.; Golestani, M.G.; Hosini, M.; Kazemian, A.

    1999-01-01

    There are some reasons for malfunction of aortic and pulmonary valve like fibrosis, calcification, and atheroma. Although, in some papers fenestration were known as a pathologic sign, but it is not generally accepted, while this matter is important in choosing suitable Homograft Heart Valve. In this paper fenestrations and its size, numbers and situation effect was studied. We collected 98 hearts, the donors died because of accident, we excluded valves with atheroma, calcification, fibrosis and unequal cusps, 91 aortic and 93 pulmonary valves were given further consideration. We classified valves according to situation, number and size of fenestration. Each valve was tested with 104 cm of non-nal saline column pressure which is equal to 76 mm Hg. Valve efficacy was detected by fluid flow assay. With study of 184 valves, 95 had no fenestration, 64 had less than 2 fenestration and 25 had more than 2 fenestration. Valve efficacy in condition of less than 2 fenestration was more than others (p <0.01). Malfunction effects of fenestration increased in larger valve and it will be decreased if their situation would be marginal (free margin of cusp). In the comparison of aortic and pulmonary valve we saw that malfunction effect of fenestration in pulmonary valve was more than aortic valve. Our experience in Immam Khomeini Homograft Valve Bank has shown that a great deal of valves is fenestrated. It seems that fenestration must be considered as a quality criterion in homograft valve preparation, especially in pulmonary and large aortic valves; but complementary studies is necessary

  5. Organic evaporator steam valve failure

    International Nuclear Information System (INIS)

    Jacobs, R.A.

    1992-01-01

    Defense Waste Processing Facility (DWPF) Technical has requested an analysis of the capacity of the Organic Evaporator (OE) condenser (OEC) be performed to determine its capability in the case where the OE steam flow control valve fails open. Calculations of the OE boilup and the OEC heat transfer coefficient indicate the OEC will have more than enough capacity to remove the heat at maximum OE boilup. In fact, the Salt Cell Vent Condenser (SCVC) should also have sufficient capacity to handle the maximum OE boilup. Therefore, it would require simultaneous loss of OEC and/or SCVC condensing capacity for the steam valve failure to cause high benzene in the Process Vessel Vent System (PVVS)

  6. Aerococcus viridans Native Valve Endocarditis

    Directory of Open Access Journals (Sweden)

    Wenwan Zhou

    2013-01-01

    Full Text Available Aerococcus viridans is an infrequent human pathogen and few cases of infective endocarditis have been reported. A case involving a 69-year-old man with colon cancer and hemicolectomy 14 years previously, without recurrence, is reported. A diagnosis of native mitral valve endocarditis was established on the basis of clinical presentation, characteristic echocardiographic findings and pathological specimen examination after urgent valve replacement. A viridans endocarditis appears to be particularly virulent, requiring a surgical approach in four of 10 cases reported and death in one of nine. Given the aggressive nature of A viridans endocarditis and the variable time to diagnosis (a few days to seven months, prompt recognition of symptoms and echocardiography, in addition to blood cultures, should be performed when symptoms persist.

  7. Transapical aortic valve implantation without angiography: proof of concept.

    Science.gov (United States)

    Ferrari, Enrico; Sulzer, Christopher; Marcucci, Carlo; Rizzo, Elena; Tozzi, Piergiorgio; von Segesser, Ludwig K

    2010-06-01

    Cardiac computed tomographic scans, coronary angiograms, and aortographies are routinely performed in transcatheter heart valve therapies. Consequently, all patients are exposed to multiple contrast injections with a following risk of nephrotoxicity and postoperative renal failure. The transapical aortic valve implantation without angiography can prevent contrast-related complications. Between November 2008 and November 2009, 30 consecutive high-risk patients (16 female, 53.3%) underwent transapical aortic valve implantation without angiography. The landmarks identification, the stent-valve positioning, and the postoperative control were routinely performed under transesophageal echocardiogram and fluoroscopic visualization without contrast injections. Mean age was 80.1 +/- 8.7 years. Mean valve gradient, aortic orifice area, and ejection fraction were 60.3 +/- 20.9 mm Hg, 0.7 +/- 0.16 cm(2), and 0.526 +/- 0.128, respectively. Risk factors were pulmonary hypertension (60%), peripheral vascular disease (70%), chronic pulmonary disease (50%), previous cardiac surgery (13.3%), and chronic renal insufficiency (40%) (mean blood creatinine and urea levels: 96.8 +/- 54 microg/dL and 8.45 +/- 5.15 mmol/L). Average European System for Cardiac Operative Risk Evaluation was 32.2 +/- 13.3%. Valve deployment in the ideal landing zone was 96.7% successful and valve embolization occurred once. Thirty-day mortality was 10% (3 patients). Causes of death were the following: intraoperative ventricular rupture (conversion to sternotomy), right ventricular failure, and bilateral pneumonia. Stroke occurred in one patient at postoperative day 9. Renal failure (postoperative mean blood creatinine and urea levels: 91.1 +/- 66.8 microg/dL and 7.27 +/- 3.45 mmol/L), myocardial infarction, and atrioventricular block were not detected. Transapical aortic valve implantation without angiography requires a short learning curve and can be performed routinely by experienced teams. Our report

  8. Bistable diverter valve in microfluidics

    Czech Academy of Sciences Publication Activity Database

    Tesař, Václav; Bandulasena, H.C.H.

    2011-01-01

    Roč. 50, č. 5 (2011), s. 1225-1233 ISSN 0723-4864 R&D Projects: GA ČR GA101/07/1499; GA AV ČR IAA200760705 Institutional research plan: CEZ:AV0Z20760514 Keywords : fluidics * bistable diverter valves * pressure-driven microfluidics Subject RIV: BK - Fluid Dynamics Impact factor: 1.735, year: 2011 http://www.springerlink.com/content/x4907p1908151522/

  9. Valve which can be remotely handled and operated, particularly for the hot area of radioactive plants. Fernhantierbare und fernbedienbare Armatur, insbesondere fuer den heissen Bereich radioaktiver Anlagen

    Energy Technology Data Exchange (ETDEWEB)

    Radovan, G.; Sandling, M.J.; Davidson, J.W.; Blaseck, K.; Hoffmeister, L.; Westendorf, H.

    1986-12-11

    The valve is built up so that one replacement part contains all the wear parts of the valve, while the non-wearing parts and particularly the valve housing are solidly connected to the pipeline. The replacement part and the valve housing are fixed on alternate sides by a clamp. The replacement part has a central screw which can be moved relative to the other components, which is supported in both axial directions in a clamp which can be opened. The replacement part can thus be pressed into the valve housing and pulled out of it.

  10. Control valve friction operational experience at Darlington NGD

    International Nuclear Information System (INIS)

    Speer, B.

    1995-01-01

    Proper installation of valve packing is an important part of ensuring that control valves operate as intended. Darlington NGD has developed a Valve Packing Program. This program combined with valve diagnostics has enabled the station to ensure that the operability of control valves is maintained after repacking. This paper outlines the process that is used for this. (author)

  11. Transcatheter aortic valve prosthesis surgically replaced 4 months after implantation

    DEFF Research Database (Denmark)

    Thyregod, Hans Gustav; Lund, Jens Teglgaard; Engstrøm, Thomas

    2010-01-01

    Transcatheter aortic valve implantation is a new and rapidly evolving treatment option for high-risk surgical patients with degenerative aortic valve stenosis. Long-term results with these new valve prostheses are lacking, and potential valve dysfunction and failure would require valve replacemen...

  12. Pivot design in bileaflet valves.

    Science.gov (United States)

    Vallana, F; Rinaldi, S; Galletti, P M; Nguyen, A; Piwnica, A

    1992-01-01

    The design criteria leading to the development of a new bileaflet valve (Sorin Bicarbon) were derived from the analysis of functional requirements, the performance of existing prostheses, and the availability of an advanced carbon coating technology (Carbofilm). The hinge is the critical element affecting fluid dynamics, durability, and thrombus formation in bileaflet valves. A comparative study of three existing models led to a new hinge design that was based on coupling two spheric surfaces with different radii of curvature (leaflet pivot and hinge recess) and obtained by electroerosion into a Carbofilm-coated metallic housing. In this valve, the point of contact moves continuously by rolling, not sliding. This minimizes friction and wear and allows uninterrupted washing of the blood exposed surfaces even during diastole (a finding established in patients using transesophageal echocardiography). Tricuspid implantation without anticoagulation in 33 sheep did not lead to thrombotic events (follow-up, 40-400 days). In the first 36 clinical implants observed for 15 months (mitral position, size 29; two unrelated deaths), the mean diastolic gradient by echo Doppler was 4 +/- 1.25 mmHg; the functional area was 3.2 +/- 0.6 cm2. No leaflet fracture and no thrombotic or embolic complications were observed clinically using a standard anticoagulant regimen.

  13. Acute Mitral Valve Dysfunction Due to Escape of Prosthetic Mechanical Leaflet and Peripheral Leaftlet Embolization.

    Science.gov (United States)

    Calik, Eyup Serhat; Limandal, Husnu Kamil; Arslan, Umit; Tort, Mehmet; Yildiz, Ziya; Bayram, Ednan; Dag, Ozgur; Kaygin, Mehmet Ali; Erkut, Bilgehan

    2015-12-14

    Leaflet escape of prosthetic valve is rare but potentially life threatening. Early diagnosis is essential on account of avoiding mortality, and emergency surgical correction is compulsory. This complication has previously been reported for both monoleaflet and bileaflet valve models. A 30-year-old man who had undergone mitral valve replacement with a bileaflet valve 8 years prior at another center was admitted with acute-onset with cardiogenic shock as an emergency case. Transthoracic echocardiograms showed acute-starting severe mitral regurgitation associated with prosthetic mitral valve. There was a suspicious finding of a single prosthetic mitral leaflet. But the problem related with the valve wasn't specifically determined. The patient underwent emergent surgery for replacement of the damaged prosthetic valves immediately. There was no tissue impingement and thrombosis, one of the two leaflets was absent, and there were no signs of endocarditis or pannus formation in the prosthetic valve. The missing leaflet could not be found within the cardiac cavity. The abdominal fluoroscopic study and plain radiography were unable to detect the escaped leaflet during surgery. The damaged valve was removed and a replacement 29 mm bileaflet mechanical valve was inserted by right lateral thoracotomy. After post-operative week one, the abdominal computed tomography scan and the ultrasound showed the escaped leaflet in the left femoral artery. Fifteen days after the surgery the escaped leaflet was removed safely from the left femoral artery and the patient made a complete recovery. The escaped leaflet showed a fracture of one of the pivot systems caused by structural failure. Early cardiac surgery should be applied because of life-threatening problems.

  14. Development of Long-Lifetime Pulsed Gas Valves for Pulsed Electric Thrusters

    Science.gov (United States)

    Burkhardt, Wendel M.; Crapuchettes, John M.; Addona, Brad M.; Polzin, Kurt A.

    2015-01-01

    even 10(exp 9) cycles is well above anything demonstrated, this lower value was selected as the design point for the present work. The valve seal must remain leak-tight throughout operation, and the body must maintain a low internal leakage at relatively high operating temperatures. The full set of design requirements used for this program are summarized in Table 1. In this work, we describe two pulsed gas valves that have been fabricated to have long lifetime and demonstrate the characteristics listed above. The first is a miniaturized, conventional electromagnet-based valve while the second is a piezoelectric-based valve design. The conventional valve, shown in Fig. 1, is opened by use of a solenoid electromagnetic actuator. When current is applied to the solenoid coil, magnetic forces pull the plunger away from the valve seat, allowing fluid to flow through the valve. Removal of electrical current permits the spring and fluid pressure to seat the plunger, halting the flow of fluid. The valve body is fabricated from 304L corrosion resistant steel (CRES) and while the parts that form the magnetic circuit are fabricated from 430 CRES. This material does not have optimum magnetic properties, but its corrosion resistance permits incorporation into a design without requiring an additional plating process. A viton O-ring compound (Parker V0884-75), selected for its mechanical strength at elevated temperatures, was used for the valve seat seal. The design was based solely on the use of analytical sizing calculations, as opposed to a more rigorous finite element analysis. While this valve is small and relatively lightweight, it does not represent a design that is optimized for mass and/or a given volume envelope. The piezoelectric valve is a "puller" valve design. Applying a voltage to the piezo crystal causes it to elongate and pull a pintle off the seat, opening the valve. The valve seal consists of the pintle with an external, spherically-formed tip fabricated from

  15. Design and performance characteristic analysis of servo valve-type water hydraulic poppet valve

    International Nuclear Information System (INIS)

    Park, Sung Hwan

    2009-01-01

    For water hydraulic system control, the flow or pressure control using high-speed solenoid valve controlled by PWM control method could be a good solution for prevention of internal leakage. However, since the PWM control of on-off valves cause extensive flow and pressure fluctuation, it is difficult to control the water hydraulic actuators precisely. In this study, the servo valve-type water hydraulic valve using proportional poppet as the main valve is designed and the performance characteristics of the servo valve-type water hydraulic valve are analyzed. Furthermore, it is demonstrated through experiments that a decline in control chamber pressure that follows the change of pilot flow is caused by the occurrence of cavitation around the proportional poppet, and that fundamental characteristics of the developed valve remain unaffected by the occurrence of cavitation

  16. 3D Printed Multimaterial Microfluidic Valve.

    Directory of Open Access Journals (Sweden)

    Steven J Keating

    Full Text Available We present a novel 3D printed multimaterial microfluidic proportional valve. The microfluidic valve is a fundamental primitive that enables the development of programmable, automated devices for controlling fluids in a precise manner. We discuss valve characterization results, as well as exploratory design variations in channel width, membrane thickness, and membrane stiffness. Compared to previous single material 3D printed valves that are stiff, these printed valves constrain fluidic deformation spatially, through combinations of stiff and flexible materials, to enable intricate geometries in an actuated, functionally graded device. Research presented marks a shift towards 3D printing multi-property programmable fluidic devices in a single step, in which integrated multimaterial valves can be used to control complex fluidic reactions for a variety of applications, including DNA assembly and analysis, continuous sampling and sensing, and soft robotics.

  17. Trans-apical aortic valve implantation in a patient with stentless valve degeneration.

    Science.gov (United States)

    Kapetanakis, Emmanouil I; MacCarthy, Philip; Monaghan, Mark; Wendler, Olaf

    2011-06-01

    Trans-apical valve-in-valve trans-catheter aortic valve implantation (TAVI) has successfully been performed in selected, high-risk patients, who suffered prosthetic degeneration after aortic valve replacement using stented xenografts. We report the case of a 79-year-old male patient who underwent one of the first successful TAVIs in a failing stentless bioprosthesis. Copyright © 2010 European Association for Cardio-Thoracic Surgery. Published by Elsevier B.V. All rights reserved.

  18. Optimal valve location in long oil pipelines

    OpenAIRE

    Grigoriev, A.; Grigorieva, N.V.

    2007-01-01

    We address the valve location problem, one of the basic problems in design of long oil pipelines. Whenever a pipeline is depressurized, the shutoff valves block the oil flow and seal the damaged part of the pipeline. Thus, the quantity of oil possibly contaminating the area around the pipeline is determined by the volume of the damaged section of the pipeline between two consecutive valves. Then, ecologic damage can be quantified by the amount of leaked oil and the environmental characteristi...

  19. Bistable (latching) solenoid actuated propellant isolation valve

    Science.gov (United States)

    Wichmann, H.; Deboi, H. H.

    1979-01-01

    The design, fabrication, assembly and test of a development configuration bistable (latching) solenoid actuated propellant isolation valve suitable for the control hydrazine and liquid fluorine to an 800 pound thrust rocket engine is described. The valve features a balanced poppet, utilizing metal bellows, a hard poppet/seat interface and a flexure support system for the internal moving components. This support system eliminates sliding surfaces, thereby rendering the valve free of self generated particles.

  20. Improved valve and dash-pot assembly

    Science.gov (United States)

    Chang, S.C.

    1985-04-23

    A dash-pot valve comprises a cylinder submerged in the fluid of a housing and have a piston attached to a plunger projecting into the path of closing movement of a pivotal valve member. A vortex chamber in said cylinder is provided with targentially directed inlets to generate vortex flow upon retraction of said plunger and effect increasing resistance against said piston to progressively retard the closing rate of said valve member toward its seat.

  1. Valve and dash-pot assembly

    Science.gov (United States)

    Chang, Shih-Chih

    1986-01-01

    A dash-pot valve comprising a cylinder submerged in the fluid of a housing and having a piston attached to a plunger projecting into the path of closing movement of a pivotal valve member. A vortex chamber in said cylinder is provided with tangentially directed inlets to generate vortex flow upon retraction of said plunger and effect increasing resistance against said piston to progressively retard the closing rate of said valve member toward its seat.

  2. Intracranial Pressure-Guided Shunt Valve Adjustments with the Miethke Sensor Reservoir.

    Science.gov (United States)

    Antes, Sebastian; Stadie, Axel; Müller, Simon; Linsler, Stefan; Breuskin, David; Oertel, Joachim

    2018-01-01

    Telemetric intracranial pressure (ICP) monitoring seems to be a promising therapy-supporting option in shunt-treated patients. Benefits become obvious when headaches are unspecific and clinical symptoms cannot be related to possible overdrainage or underdrainage. In this study, we evaluated a new telemetric device to individually adjust shunt valves according to ICP measurements. Between December 2015 and November 2016, 25 patients with suspected suboptimal shunt valve settings underwent insertion of a telemetric ICP sensor (Sensor Reservoir; Christoph Miethke, Potsdam, Germany). Over a 1-year period, a total of 183 telemetric ICP measurements and 85 shunt valve adjustments were carried out. Retrospective statistic analyses focused on valve adjustments, ICP values, and clinical outcomes. ICP-guided valve adjustments positively changed the clinical state in 18 out of 25 patients. Clinical improvement over time was associated with significant changes of the valve settings and ICP values. Interestingly, a therapeutically normalized ICP profile was not automatically associated with clinical improvement. The Sensor Reservoir is an important and valuable tool for shunt-treated patients suffering from drainage-related problems. The possibility to simultaneously recognize and solve shunt problems represents the decisive advantage. Nevertheless, measurements with the Sensor Reservoir did not allow for the determination of default valve settings or universal target ICP values. Copyright © 2017 Elsevier Inc. All rights reserved.

  3. A 3D velocimetry study of the flow through prosthetic heart valves

    Science.gov (United States)

    Ledesma, R.; Zenit, R.; Pulos, G.; Sanchez, E.; Juarez, A.

    2006-11-01

    Blood damage commonly appears in medical valve prothesis. It is a mayor concern for the designers and surgeons. It is well known that this damage and other complications result from the modified fluid dynamics through the replacement valve. To evaluate the performance of prosthetic heart valves, it is necessary to study the flow through them. To conduct this study , we have built a flow channel that emulates cardiac conditions and allows optical access such that a 3D-PIV velocimetry system could be used. The experiments are aimed to reconstruct the downstream structure of the flow through a mechanical and a bio-material tricuspid heart valve prothesis. Preliminary results show that the observed coherent structures can be related with haemolysis and trombosis, illnesses commonly found in valve prothesis recipients. The mean flow, the levels of strain rate and the turbulence intensity generated by the valves can also be directly related to blood damage. In general, bio-material made valves tend to reduce these complications.

  4. Fast Flux Test Facility primary sodium valves

    International Nuclear Information System (INIS)

    Rabe, G.B.; Ezra, B.C.

    1977-01-01

    The design and development of the valves used in the primary sodium coolant loop of the Fast Flux Test Facility is described. One tilting-disk check valve is used in the cold leg of the coolant loop. It is designed to limit flow reversal in the loop while maintaining a low pressure drop during forward flow. Two isolation valves are used in each coolant loop--one in the cold leg and one in the hot leg. They are of the motor-operated swinging-gate type. The design, analysis, and testing programs undertaken to develop and qualify these valves are described

  5. Advantages of butterfly valves for power plants

    International Nuclear Information System (INIS)

    Lapadat, J.T.

    1977-01-01

    Butterfly valves are increasingly used in nuclear power plants. They are used in CANDU reactors for class 2 and 3 service, to provide emergency and tight shutoff valves for all inlets and outlets of heat exchangers and all calandria penetrations. Guidelines for meeting nuclear power plant valve specifications are set out in ASME Section 3, Nuclear Power Plant Components. Some details of materials of construction, type of actuator, etc., for various classes of nuclear service are tabulated in the present article. The 'fishtail' butterfly valve is an improved design with reduced drag, as is illustrated and explained. (N.D.H.)

  6. Additively Manufactured Main Fuel Valve Housing

    Science.gov (United States)

    Eddleman, David; Richard, Jim

    2015-01-01

    Selective Laser Melting (SLM) was utilized to fabricate a liquid hydrogen valve housing typical of those found in rocket engines and main propulsion systems. The SLM process allowed for a valve geometry that would be difficult, if not impossible to fabricate by traditional means. Several valve bodies were built by different SLM suppliers and assembled with valve internals. The assemblies were then tested with liquid nitrogen and operated as desired. One unit was also burst tested and sectioned for materials analysis. The design, test results, and planned testing are presented herein.

  7. Safety valve opening and closing operation monitor

    International Nuclear Information System (INIS)

    Kodama, Kunio; Takeshima, Ikuo; Takahashi, Kiyokazu.

    1981-01-01

    Purpose: To enable the detection of the closing of a safety valve when the internal pressure in a BWR type reactor is a value which will close the safety valve, by inputting signals from a pressure detecting device mounted directly at a reactor vessel and a safety valve discharge pressure detecting device to an AND logic circuit. Constitution: A safety valve monitor is formed of a pressure switch mounted at a reactor pressure vessel, a pressure switch mounted at the exhaust pipe of the escape safety valve and a logic circuit and the lide. When the input pressure of the safety valve is raised so that the valve and the pressure switch mounted at the exhaust pipe are operated, an alarm is indicated, and the operation of the pressure switch mounted at a pressure vessel is eliminated. If the safety valve is not reclosed when the vessel pressure is decreased lower than the pressure at which it is to be reclosed after the safety valve is operated, an alarm is generated by the logic circuit since both the pressure switches are operated. (Sekiya, K.)

  8. Outcome of pregnancy in women after pulmonary autograft valve replacement for congenital aortic valve disease

    NARCIS (Netherlands)

    Yap, Sing-Chien; Drenthen, Willem; Pieper, Petronella G.; Moons, Philip; Mulder, Barbara J. M.; Klieverik, Loes M.; Vliegen, Hubert W.; van Dijk, Arie P. J.; Meijboom, Folkert J.; Roos-Hesselink, Jolien W.

    Background and aim of the study: The pulmonary autograft has been recommended as the valve of choice for aortic valve replacement (AVR) in young women contemplating pregnancy. However, current information on maternal and perinatal outcome of pregnancy in women with pulmonary autograft valve

  9. Outcome of pregnancy in women after pulmonary autograft valve replacement for congenital aortic valve disease

    NARCIS (Netherlands)

    Yap, Sing-Chien; Drenthen, Willem; Pieper, Petronella G.; Moons, Philip; Mulder, Barbara J. M.; Klieverik, Loes M.; Vliegen, Hubert W.; van Dijk, Arie P. J.; Meijboom, Folkert J.; Roos-Hesselink, Jolien W.

    2007-01-01

    Background and aim of the study: The pulmonary autograft has been recommended as the valve of choice for aortic valve replacement (AVR) in young women contemplating pregnancy. However, current information on maternal and perinatal outcome of pregnancy in women with pulmonary autograft valve

  10. Outcome of pregnancy in women after pulmonary autograft valve replacement for congenital aortic valve disease.

    NARCIS (Netherlands)

    Yap, S.C.; Drenthen, W.; Pieper, P.G.; Moons, P.; Mulder, B.J.M.; Klieverik, L.M.; Vliegen, H.W.; Dijk, A.P.J. van; Meijboom, F.J.; Roos-Hesselink, J.W.

    2007-01-01

    BACKGROUND AND AIM OF THE STUDY: The pulmonary autograft has been recommended as the valve of choice for aortic valve replacement (AVR) in young women contemplating pregnancy. However, current information on maternal and perinatal outcome of pregnancy in women with pulmonary autograft valve

  11. Transapical JenaValve in a patient with mechanical mitral valve prosthesis.

    LENUS (Irish Health Repository)

    O' Sullivan, Katie E

    2014-01-29

    We report the first case of transcatheter aortic valve replacement implantation using JenaValve™ in a patient with mechanical mitral valve prosthesis. We believe that the design features of this valve may be particularly suited for use in this setting. © 2014 Wiley Periodicals, Inc.

  12. Safety valve including a hydraulic brake and hydraulic brake that could be fitted into a valve

    International Nuclear Information System (INIS)

    Chabat-Courrede, Jean.

    1981-01-01

    Making of a safety valve that can be fitted to a containment vessel filled with a non compressible fluid, such as the water system of a nuclear power station. It includes a hydraulic brake located between the valve and the elastic means, close to the valve which completely suppresses the high frequency oscillations of the equipment [fr

  13. Small valve area index: its influence on early mortality after mitral valve replacement

    NARCIS (Netherlands)

    Yazdanbakhsh, A. P.; van den Brink, R. B.; Dekker, Egbert; de Mol, B. A.

    2000-01-01

    OBJECTIVE: To test the hypothesis that mitral valve prosthesis-patient mismatch increases postoperative mortality. METHODS AND RESULTS: The effect of mitral valve prosthesis-patient mismatch on survival in a cohort of consecutive patients after mitral valve replacement with a mechanical prosthesis

  14. Impact of bicuspid aortic valve on complications and death in infective endocarditis of native aortic valves.

    Science.gov (United States)

    Kahveci, Gokhan; Bayrak, Fatih; Pala, Selcuk; Mutlu, Bulent

    2009-01-01

    We retrospectively investigated the impact of bicuspid aortic valve on the prognosis of patients who had definite infective endocarditis of the native aortic valve.Of 51 patients, a bicuspid aortic valve was present in 22 (43%); the other 29 had tricuspid aortic valves. On average, the patients who had bicuspid valves were younger than those who had tricuspid valves. Patients with a tricuspid valve had larger left atrial diameters and were more likely to have severe mitral regurgitation.Periannular complications, which we detected in 19 patients (37%), were much more common in the patients who had a bicuspid valve (64% vs 17%, P = 0.001). The presence of a bicuspid valve was the only significant independent predictor of periannular complications. The in-hospital mortality rate in the bicuspid group was lower than that in the tricuspid group; however, this figure did not reach statistical significance (9% vs 24%, P = 0.15). In multivariate analysis, left atrial diameter was the only independent predictor associated with an increased risk of death (hazard ratio, 2.19; 95% confidence interval, 1.1-4.5; P = 0.031).In our study, patients with infective endocarditis in a bicuspid aortic valve were younger and had a higher incidence of periannular complications. Although a worse prognosis has been reported previously, we found that infective endocarditis in a native bicuspid aortic valve is not likely to increase the risk of death in comparison with infective endocarditis in native tricuspid aortic valves.

  15. Initial data collection efforts of CREDO. Sodium valve failures

    International Nuclear Information System (INIS)

    Bott, T.F.; Haas, P.M.

    1978-01-01

    The Centralized Reliability Data organisation (CREDO) has been established at Oak Ridge National Laboratory to define, develop, and maintain a reliability data analysis center for use in advanced reactor safety and licensing. Its primary functions are collection, reduction, evaluation, storage, retrieval, and dissemination of reliability/maintainability data. Data-collection efforts have been initiated at several test loops, at the Experimental Breeder Reactor-II and at the Fast Flux Test Facility. Top priority is being given to collection data on safety and safety-related systems, primarily for sodium-cooled reactors. Sufficient operating time has been accumulated on sodium valves at test facilities to provide quantitative estimates of reliability characteristics with a reasonable degree of confidence. Sodium-valve failures have been categorized according to seat design, size, seal type, and actuator type. Attempts have been made to establish the variation of failure rate with time and duty. Estimates of failure rates for sodium valves have been compared to those for water valves and appear to be of the same order of magnitude. (author)

  16. Hydraulic engine valve actuation system including independent feedback control

    Science.gov (United States)

    Marriott, Craig D

    2013-06-04

    A hydraulic valve actuation assembly may include a housing, a piston, a supply control valve, a closing control valve, and an opening control valve. The housing may define a first fluid chamber, a second fluid chamber, and a third fluid chamber. The piston may be axially secured to an engine valve and located within the first, second and third fluid chambers. The supply control valve may control a hydraulic fluid supply to the piston. The closing control valve may be located between the supply control valve and the second fluid chamber and may control fluid flow from the second fluid chamber to the supply control valve. The opening control valve may be located between the supply control valve and the second fluid chamber and may control fluid flow from the supply control valve to the second fluid chamber.

  17. A shut-off valve for flexible tubing

    Science.gov (United States)

    Reyburn, W. W.

    1972-01-01

    Design of light weight valve for flexible tubing is described. Valve is hand operated and provides positive sealing in normally closed position. Diagram is provided to show construction of valve. Principles of operation are explained.

  18. North American trial results at 1 year with the Sorin Freedom SOLO pericardial aortic valve.

    Science.gov (United States)

    Heimansohn, David; Roselli, Eric E; Thourani, Vinod H; Wang, Shaohua; Voisine, Pierre; Ye, Jian; Dabir, Reza; Moon, Michael

    2016-02-01

    A North American prospective, 15-centre Food and Drug Administration (FDA) valve trial was designed to assess the safety and effectiveness of the Freedom SOLO stentless pericardial aortic valve in the treatment of surgical aortic valve disease. Beginning in 2010, 251 patients (mean: 74.7 ± 7.5 years), were recruited in the Freedom SOLO aortic valve trial. One hundred eighty-nine patients have been followed for at least 1 year and are the basis for this review. Preoperatively, 54% of patients had NYHA functional class III or IV symptoms, and the majority of patients had a normal ejection fraction (EF) (median EF = 61%). Concomitant procedures were performed in 61.9% of patients, with coronary artery bypass grafting (CABG) (48.7%) being the most common followed by a MAZE procedure (13.7%). Reoperations were performed in 8.5% of patients in the study. The entire cohort of 251 patients enrolled had 7 deaths prior to 30 days, 2 of which were valve-related (aspiration pneumonia and sudden death) and 5 were not valve-related. There were 11 deaths after 30 days, 1 valve-related (unknown cardiac death) and 10 not valve-related. Five valves were explanted, 3 early (endocarditis, acute insufficiency and possible root dissection) and 2 late (endocarditis). Thirty-day adverse events include arrhythmias requiring permanent pacemaker (4.2%), thromboembolic events (3.7%) and thrombocytopenia (7.4%). One-year follow-up of all 189 patients demonstrated mean gradients for valve sizes 19, 21, 23, 25 and 27 mm of 11.7, 7.8, 6.3, 4.6 and 5.0 mmHg, respectively. Effective orifice areas for the same valve sizes were 1.2, 1.3, 1.6, 1.8 and 1.9 cm(2), respectively. Ninety-six percent of patients (181/189) were in NYHA class I or II at the 1-year follow-up. The Freedom SOLO stentless pericardial aortic valve demonstrated excellent haemodynamics and a good safety profile out to the 1 year of follow-up. © The Author 2015. Published by Oxford University Press on behalf of the European

  19. Preventative valve-sparing aortic root replacement and pregnancy outcome in Marfan syndrome.

    Science.gov (United States)

    Sokol, Vesna; Zlopasa, Gordan; Herman, Mislav; Planinić, Pavao; Micevska, Ana

    2012-06-01

    In Marfan syndrome, with dilatation of the aortic root secondary to an underlying connective tissue defect, pregnancy can cause hemodynamic stress leading to the development of an aortic aneurysm and even a fatal aortic dissection. In the presence of existing aortic root enlargement and a family history of aortic dissection, preventative elective surgery is suggested. Aortic root replacement with or without a valve-sparing procedure is superior to total aortic root replacement with prosthetic valve/tube graft. It provides excellent survival with low rates of aortic - valve related complications.

  20. Removable butterfly valve, especially for polluting and/or dangerous fluids

    International Nuclear Information System (INIS)

    Garrigues, J.C.

    1984-01-01

    This invention relates to a valve for use in systems carrying polluting, corrosive or dangerous fluids requiring the use of biological protection. The facilities concerned are those in which fluids, mainly polluting, corrosive or dangerous liquids requiring the use of various types of biological protection, are handled. This is particularly so for nuclear installations in which the equipment is surrounded by protective shields which stop the radiation and prevent radioactive gases and aerosols from spreading. The invention proposes for the present valve a high-safety leaktightness system which respects the most specifications and standards and which suppresses any dead volume inside the valve [fr

  1. Successful salvage treatment of native valve Enterococcus faecalis infective endocarditis with telavancin: two case reports.

    Science.gov (United States)

    Thompson, Mickala M; Hassoun, Ali

    2017-07-01

    Infective endocarditis (IE) one-year mortality rates approach 40%. Here, we report two native valve Enterococcus faecalis IE cases in patients successfully treated with telavancin. An 88-year-old with mitral valve endocarditis and a penicillin allergy, initially treated with intravenous vancomycin, was switched to telavancin. A 69-year-old, who previously received amoxicillin and intravenous vancomycin for presumed enterococcal bacteraemia, was diagnosed with dual valve endocarditis for which he received telavancin. Both received six weeks of telavancin. Neither had telavancin-related adverse events, evidence of infection at six months, nor required telavancin dosing adjustments. Documented use of novel treatments for serious enterococcal infections is needed.

  2. Analysis of pressure losses in the diffuser of a control valve

    Science.gov (United States)

    Turecký, Petr; Mrózek, Lukáš; Tajč, Ladislav; Kolovratník, Michal

    The pressure loss in the diffuser of a control valve is evaluated by using CFD computations. Pressure ratios and lifts of a cone for the recommended flow characteristics of an experimental turbine are considered. The pressure loss in a valve is compared with the pressure loss in a nozzle, i.e. the embodiment of the valve without a cone. Computations are carried out for the same mass flow. Velocity profiles are evaluated in both versions of computations. Comparison of computed pressure losses, with the loss evaluated by using relations for diffusers with the ideal velocity conditions in the input cross-section, is carried out.

  3. Analysis of pressure losses in the diffuser of a control valve

    Directory of Open Access Journals (Sweden)

    Turecký Petr

    2017-01-01

    Full Text Available The pressure loss in the diffuser of a control valve is evaluated by using CFD computations. Pressure ratios and lifts of a cone for the recommended flow characteristics of an experimental turbine are considered. The pressure loss in a valve is compared with the pressure loss in a nozzle, i.e. the embodiment of the valve without a cone. Computations are carried out for the same mass flow. Velocity profiles are evaluated in both versions of computations. Comparison of computed pressure losses, with the loss evaluated by using relations for diffusers with the ideal velocity conditions in the input cross-section, is carried out.

  4. Outcomes of Aortic Valve-Sparing Operations in Marfan Syndrome.

    Science.gov (United States)

    David, Tirone E; David, Carolyn M; Manlhiot, Cedric; Colman, Jack; Crean, Andrew M; Bradley, Timothy

    2015-09-29

    In many cardiac units, aortic valve-sparing operations have become the preferred surgical procedure to treat aortic root aneurysm in patients with Marfan syndrome, based on relatively short-term outcomes. This study examined the long-term outcomes of aortic valve-sparing operations in patients with Marfan syndrome. All patients with Marfan syndrome operated on for aortic root aneurysm from 1988 through 2012 were followed prospectively for a median of 10 years. Follow-up was 100% complete. Time-to-event analyses were calculated using the Kaplan-Meier method with log-rank test for comparisons. A total of 146 patients with Marfan syndrome had aortic valve-sparing operations. Reimplantation of the aortic valve was performed in 121 and remodeling of the aortic root was performed in 25 patients. Mean age was 35.7 ± 11.4 years and two-thirds were men. Nine patients had acute, 2 had chronic type A, and 3 had chronic type B aortic dissections before surgery. There were 1 operative and 6 late deaths, 5 caused by complications of dissections. Mortality rate at 15 years was 6.8 ± 2.9%, higher than the general population matched for age and sex. Five patients required reoperation on the aortic valve: 2 for endocarditis and 3 for aortic insufficiency. Three patients developed severe, 4 moderate, and 3 mild-to-moderate aortic insufficiency. Rate of aortic insufficiency at 15 years was 7.9 ± 3.3%, lower after reimplantation than remodeling. Nine patients developed new distal aortic dissections during follow-up. Rate of dissection at 15 years was 16.5 ± 3.4%. Aortic valve-sparing operations in patients with Marfan syndrome were associated with low rates of valve-related complications in long-term follow-up. Residual and new aortic dissections were the leading cause of death. Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  5. Simulation of Valve Operation for Flow Interrupt Test in Nuclear Power Plant

    International Nuclear Information System (INIS)

    Kim, Jae Hyung; Shin, Dae Yong; Shin, Dong Woo; Kim, Charn Jung; Lee, Jung Hee

    2012-01-01

    The valve used in nuclear power plant must be qualified for the function according to the KEPIC MF. The test valve must be selected by shape and size, which is given by KEPIC MF. In the functional test, the mathematical model for the valve operation is needed. The mathematical model must be verified by the test, whose method and procedure is defined in KEPIC MF. The lack of analytical technique has lead to the poor mathematical model, with which the functional test for the big valve is impossible with analytical method. Especially, the tank and rupture disk in the flow test is not considered and the result of the analysis is so different to the real one. In these days, the 3D model for the flow interrupt test makes more accurate analysis. And no facility about functional test reduces the research will for the nuclear power plant valve. For this problem, the test facility for the functional test of the valve and pump in nuclear power plant has been made until 2012. With the test facility, the research project related the valve were initiated in KIMM( Korea Institute of Machinery and Materials). And the joint project to SNU(Seoul National University) has been going on the numerical analysis for the valve in nuclear power plant. Using the commercial software and user subroutine, UDF, the co-simulation with multi-body dynamic and fluid flow analysis and the addition of tank and rupture disk to the user subroutine make possible to simulate the flow interrupt test numerically. This is not simple and regular analysis, which was introduced in user subroutine. In order to simulate the real situation, the engineering work, related mathematical model, and the programming in the user subroutine are needed. This study is on the making the mathematical model for the functional test of the valve in nuclear power plan. The functional test is the real test procedure and defined in KEPIC MF

  6. Ahmed glaucoma valve in children: A review☆

    Science.gov (United States)

    Nassiri, Nariman; Nouri-Mahdavi, Kouros; Coleman, Anne L.

    2011-01-01

    Pediatric glaucoma is potentially a blinding disease. Although goniotomy and trabeculotomy are associated with good early success rates, eventually 20% of these procedures fail and many children will require additional surgery to control the IOP in the long-term. In this review, we reported that adequate IOP control can be achieved with the placement of Ahmed glaucoma valve and can last 5 or more years. However, most patients will need one or more glaucoma medications at some point after surgery. In addition, the implants may be associated with pupillary irregularities, lenticular opacification as well as tube-related complications, particularly in the first year of life, as the globe is enlarging with age. PMID:23960945

  7. Resurgery for recurrent heart valve diseases

    Directory of Open Access Journals (Sweden)

    Chong-lei REN

    2017-02-01

    Full Text Available Objective To summarize the experience with resurgery for recurrent valvular heart diseases. Methods From June 2004 to June 2015, 28 patients (15 males and 13 females with ages ranging from 44 to 67 years (55.6±6.5 years with recurrent heart valve disease underwent resurgery. The reasons for resurgery included perivalvular leakage (7 cases, bioprosthetic valve decline (6 cases in mitral valve and 3 in tricuspid valve, mechanical prostheses dysfunction (2cases, infective endocarditis after valve replacement (2 cases, restenosis of repaired native valve (1 case, and severe tricuspid insufficiency after left-side valve surgery (7 cases. Resurgery included mitral valve replacement in 18 patients and tricuspid valve replacement in 10. All the patients underwent third or fourth or even fifth cardiac surgery for valve replacement. Results There were 2 hospital deaths with a mortality of 7.1% (2/28. The main causes of early-stage deaths were low cardiac output syndrome. The main postoperative complications were respiratory failure in 3, low cardiac output syndrome in 2, reexploration for bleeding in 2 and serious infectious shock in 1. All the patients were found with the great improvement in heart function and the re-implanted prostheses worked well during follow-up. Conclusions Although resurgery for recurrent heart valve disease poses a continuing challenge to cardiac surgeon, it could be performed with the satisfactory results. The keys to a successful cardiac resurgery include appropriate operational timing, refined surgical technique and reasonable perioperative managements. DOI: 10.11855/j.issn.0577-7402.2017.01.11

  8. Multidetector computed tomography sizing of bioprosthetic valves: guidelines for measurement and implications for valve-in-valve therapies

    International Nuclear Information System (INIS)

    Rajani, R.; Attia, R.; Condemi, F.; Webb, J.; Woodburn, P.; Hodson, D.; Nair, A.; Preston, R.; Razavi, R.; Bapat, V.N.

    2016-01-01

    Aim: To describe a technique for bioprosthetic multidetector computed tomography (MDCT) sizing and to compare MDCT-derived values against manufacturer-provided sizing. Materials and methods: Fourteen bioprosthetic stented valves commonly used in the aortic valve position were evaluated using a Philips 256 MDCT system. All valves were scanned using a dedicated cardiac CT protocol with a four-channel electrocardiography (ECG) simulator. Measurements were made of major and minor axes and the area and perimeter of the internal stent using varying reconstruction kernels and window settings. Measurements derived from MDCT (MDCT ID) were compared against the stent internal diameter (Stent ID) as provided by the valve manufacturer and the True ID (Stent ID + insertion of leaflets). All data were collected and analysed using SPSS for Mac (version 21). Results: The mean difference between the MDCT ID and Stent ID was 0.6±1.9 mm (r=0.649, p=0.012) and between MDCT ID and True ID 2.1±2 mm (r=0.71, p=0.005). There was no difference in the major (p=0.90), minor (p=0.87), area (p=0.92), or perimeter (p=0.92) measurements when sharp, standard, and detailed stent kernels were used. Similarly, the measurements remained consistent across differing windowing levels. Conclusion: Bioprosthetic stented valves may be reliably sized using MDCT in patients requiring valve-in-valve (VIV) interventions where the valve type and size are unknown. In these cases, clinicians should be aware that MDCT has a tendency to overestimate the True ID size. - Highlights: • Cardiac CT is likely to be ideally suited for bioprosthetic aortic valve sizing for valve in valve procedures. • We compared MDCT sizing for 14 varying bioprosthetic aortic valves across varying window settings and reconstruction kernels. • We provide “normal” MDCT sizing for varying valves and show their relationship to surgical sizing. • Bioprosthetic valves may be reliably sized by MDCT but require adjustment owing to

  9. Calcific Aortic Valve Disease Is Associated with Layer-Specific Alterations in Collagen Architecture.

    Directory of Open Access Journals (Sweden)

    Heather N Hutson

    Full Text Available Disorganization of the valve extracellular matrix (ECM is a hallmark of calcific aortic valve disease (CAVD. However, while microarchitectural features of the ECM can strongly influence the biological and mechanical behavior of tissues, little is known about the ECM microarchitecture in CAVD. In this work, we apply advanced imaging techniques to quantify spatially heterogeneous changes in collagen microarchitecture in CAVD. Human aortic valves were obtained from individuals between 50 and 75 years old with no evidence of valvular disease (healthy and individuals who underwent valve replacement surgery due to severe stenosis (diseased. Second Harmonic Generation microscopy and subsequent image quantification revealed layer-specific changes in fiber characteristics in healthy and diseased valves. Specifically, the majority of collagen fiber changes in CAVD were found to occur in the spongiosa, where collagen fiber number increased by over 2-fold, and fiber width and density also significantly increased. Relatively few fibrillar changes occurred in the fibrosa in CAVD, where fibers became significantly shorter, but did not otherwise change in terms of number, width, density, or alignment. Immunohistochemical staining for lysyl oxidase showed localized increased expression in the diseased fibrosa. These findings reveal a more complex picture of valvular collagen enrichment and arrangement in CAVD than has previously been described using traditional analysis methods. Changes in fiber architecture may play a role in regulating the pathobiological events and mechanical properties of valves during CAVD. Additionally, characterization of the ECM microarchitecture can inform the design of fibrous scaffolds for heart valve tissue engineering.

  10. Numerical investigation on cavitation in pressure relief valve for coal liquefaction

    International Nuclear Information System (INIS)

    Ou, G F; Li, W Z; Xiao, D H; Zheng, Z J; Dou, H S; Wang, C

    2015-01-01

    The pressure relief valve for regulating the level of the high-pressure separator works under a pressure difference up to 15 MPa in the temperature of 415 °C. Severe cavitation erosion and particle impact lead to the valve disc's mass loss. In this paper, three-dimensional turbulent cavitating flows in the pressure relief valve are numerically simulated to reveal the mechanism of mass loss at valve disc. The RNG k-ε turbulence model and the mixture model with a mass transfer for cavitation are employed to simulate the cavitating flow in the pressure relief valve. The result shows that there is phase change in the pressure relief process and cavitation bubbles would be transported by high-velocity backflow to the head of valve disc. For the local pressure higher than the saturated vapor pressure, the bubbles collapse at the head of disc and cavitation erosion is formed at the head of the disc. By comparing the cases of opening of 40%, 50%, and 60%, backflow velocity and cavitation region in front of the disc decrease with the opening increase. Therefore, during the actual operation, the pressure relief valve should be kept to a relatively large opening

  11. Study of check valve slamming in a BWR feedwater system following a postulated pipe break

    International Nuclear Information System (INIS)

    Safwat, H.H.; Arastu, A.H.; Norman, A.

    1985-01-01

    This study deals with a swing check valve slamming due to a break at relatively short distance from the valve. Under this situation, substantial flashing occurs near the valve and the result of the study are subject to what is believed to be a conservative simplifying assumption, i.e., the hydrodynamic moment acting on the valve during the transient is represented by resultant moment due to the pressure differential across the valve. It is believed that vapor voids forming at the valve would actually reduce the disk impact velocities in comparison to those predicted under this simplifying assumption. A technique used to represent a double-ended break through hypothetical valves may have some influence on the results particularly for long break opening times. The study has yielded good insight to help understand the complex problem. The study has focused on some parameters and the reader may raise questions on the effects of other parameters. Nevertheless, the present study underlines the complexity facing analysts dealing with this transient using analytical methods. Though some experimental data are available, the authors believe that an experimental study (recognizing the complexity of the experimental setup and instrumentation), would be quite useful. It can provide answers to questions facing analysts dealing with this problem and thus avoid unnecessary conservatisms due to uncertainties in input data

  12. Transcatheter valve-in-valve implantation due to severe aortic regurgitation in a degenerated aortic homograft

    DEFF Research Database (Denmark)

    Olsen, Lene Kjaer; Engstrøm, Thomas; Søndergaard, Lars

    2009-01-01

    Transcatheter aortic valve implantation (TAVI) in severe aortic stenosis has proven to be a feasible and effective treatment modality for inoperable patients. Until now, neither aortic regurgitation nor degenerated bioprostheses has been an indication for TAVI. However, this article reports...... a successful valve-in-valve implantation of a CoreValve aortic valve prosthesis through the right subclavian artery in a case of severe aortic regurgitation within a degenerated aortic homograft. The case exemplifies the possibilities of expanding the indications for TAVI, as well as other vascular access...

  13. Gasoline New Timing and Flux Adjustable Rotary Valve Design (Hereinafter: Rotary Valve

    Directory of Open Access Journals (Sweden)

    Du huiqi

    2016-01-01

    Full Text Available Conventional gasoline engine with an umbrella valve control cylinder intake and exhaust, in order to achieve sealing effect, the valve is driven by the spring force; at the same time, when the cam opens the valve to overcome the spring force acting. Sealing the better, the more power consumed in the engine mechanical losses, the valve mechanism consumes about 30%, which is not a small loss! This article describes a new type of rotary valve is to significantly reduce mechanical losses, so as to achieve energy saving purposes.

  14. High Reliability Cryogenic Piezoelectric Valve Actuator, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — Cryogenic fluid valves are subject to harsh exposure and actuators to drive these valves require robust performance and high reliability. DSM's piezoelectric...

  15. Dikkers Valves for nuclear industry

    International Nuclear Information System (INIS)

    Anon.

    1975-01-01

    Most countries have adopted the ASME Boiler and Pressure Vessel Code Section III, as the basis of their national requirements for licensing nuclear components. This Code gives clear directives for defining design requirements coupled with a controlled manufacturing system. It has always been and still is the policy of Dikkers to manufacture high-quality products. Dikkers manufacture nuclear products in accordance with this Code, Section III; indeed many features exceed these minimum requirements. At the Nuclex Exhibition in Basel, Dikkers Valves BV will exhibit its main products for use in nuclear power plants. (Auth.)

  16. Cylinder valve packing nut studies

    Energy Technology Data Exchange (ETDEWEB)

    Blue, S.C. [Martin Marietta Energy Systems, Inc., Paducah, KY (United States)

    1991-12-31

    The design, manufacture, and use of cylinder valve packing nuts have been studied to improve their resistance to failure from stress corrosion cracking. Stress frozen photoelastic models have been analyzed to measure the stress concentrations at observed points of failure. The load effects induced by assembly torque and thermal expansion of stem packing were observed by strain gaging nuts. The effects of finishing operations and heat treatment were studied by the strain gage hole boring and X-ray methods. Modifications of manufacturing and operation practices are reducing the frequency of stress corrosion failures.

  17. Valve Corporation: Composing Internal Markets

    OpenAIRE

    Todd R. Zenger

    2015-01-01

    Discussions of the Valve Corporation are always enlightening. The skeptic wonders how much is rhetoric and recruiting ploy and how much is real. Is there clear evidence that this organizational design actually works – that it is efficient in this setting? While revenues per employee are quite remarkable, cause and effect are unclear. Is “boss-less-ness” the cause of high sales per employee or simply the result of high sales per employee, fueled from earlier success? The same question could be...

  18. Clinical Implication of Aortic Wall Biopsy in Aortic Valve Disease with Bicuspid Valve Pathology

    Directory of Open Access Journals (Sweden)

    Yong Han Kim

    2016-12-01

    Full Text Available Background: Although unique aortic pathology related to bicuspid aortic valve (BAV has been previously reported, clinical implications of BAV to aortopathy risk have yet to be investigated. We looked for potential differences in matrix protein expressions in the aortic wall in BAV patients. Methods: Aorta specimens were obtained from 31 patients: BAV group (n=27, tricuspid aortic valve (TAV group (n=4. The BAV group was categorized into three subgroups: left coronary sinus-right coronary sinus (R+L group; n=13, 42%, right coronary sinus-non-coronary sinus (R+N group; n=8, 26%, and anteroposterior (AP group; n=6, 19%. We analyzed the expression of endothelial nitric oxide synthase (eNOS, matrix metalloproteinase (MMP-9, and tissue inhibitor of matrix metalloproteinase (TIMP-2. Results: Based on the mean value of the control group, BAV group showed decreased expression of eNOS in 72.7% of patients, increased MMP-9 in 82.3%, and decreased TIMP in 79.2%. There was a higher tendency for aortopathy in the BAV group: eNOS (BAV:TAV= 53%±7%:57%±11%, MMP-9 (BAV:TAV=48%±10%:38%±1%. The AP group showed lower expression of eNOS than the fusion (R+L, R+N group did; 48%±5% vs. 55%±7% (p=0.081. Conclusion: Not all patients with BAV had expression of aortopathy; however, for patients who had a suspicious form of bicuspid valve, aortic wall biopsy could be valuable to signify the presence of aortopathy.

  19. Transcatheter aortic valve-in-valve treatment of degenerative stentless supra-annular Freedom Solo valves: A single centre experience.

    Science.gov (United States)

    Cockburn, James; Dooley, Maureen; Parker, Jessica; Hill, Andrew; Hutchinson, Nevil; de Belder, Adam; Trivedi, Uday; Hildick-Smith, David

    2017-02-15

    Redo surgery for degenerative bioprosthetic aortic valves is associated with significant morbidity and mortality. Report results of valve-in-valve therapy (ViV-TAVI) in failed supra-annular stentless Freedom Solo (FS) bioprostheses, which are the highest risk for coronary occlusion. Six patients with FS valves (mean age 78.5 years, 50% males). Five had valvular restenosis (peak gradient 87.2 mm Hg, valve area 0.63 cm 2 ), one had severe regurgitation (AR). Median time to failure was 7 years. Patients were high risk (mean STS/Logistic EuroScore 10.6 15.8, respectively). FS valves ranged from 21 to 25 mm. Successful ViV-TAVI was achieved in 4/6 patients (67%). Of the unsuccessful cases, (patient 1 and 2 of series) patient 1 underwent BAV with simultaneous aortography which revealed left main stem occlusion. The procedure was stopped and the patient went forward for repeat surgery. Patient 2 underwent successful ViV-TAVI with a 26-mm CoreValve with a guide catheter in the left main, but on removal coronary obstruction occurred, necessitating valve snaring into the aorta. Among the successful cases, (patients 3, 4, 5, 6) the TAVIs used were CoreValve Evolut R 23 mm (n = 3), and Lotus 23 mm (n = 1). In the successful cases the peak gradient fell from 83.0 to 38.3 mm Hg. No patient was left with >1+ AR. One patient had a stroke on Day 2, with full neurological recovery. Two patients underwent semi-elective pacing for LBBB and PR >280 ms. ViV-TAVI in stentless Freedom Solo valves is high risk. The risk of coronary occlusion is high. The smallest possible prosthesis (1:1 sizing) should be used, and strategies to protect the coronary vessels must be considered. © 2016 Wiley Periodicals, Inc. © 2016 Wiley Periodicals, Inc.

  20. Structural valve deterioration in a starr-edwards mitral caged-disk valve prosthesis.

    Science.gov (United States)

    Aoyagi, Shigeaki; Tayama, Kei-Ichiro; Okazaki, Teiji; Shintani, Yusuke; Kono, Michitaka; Wada, Kumiko; Kosuga, Ken-Ichi; Mori, Ryusuke; Tanaka, Hiroyuki

    2013-01-01

    The durability of the Starr-Edwards (SE) mitral caged-disk valve, model 6520, is not clearly known, and structural valve deterioration in the SE disk valve is very rare. Replacement of the SE mitral disk valve was performed in 7 patients 23-40 years after implantation. Macroscopic examination of the removed disk valves showed no structural abnormalities in 3 patients, in whom the disk valves were removed at valves excised >36 years after implantation in 4 patients. Disk fracture, a longitudinal split in the disk along its circumference at the site of incorporation of the titanium ring, was detected in the valves removed 36 and 40 years after implantation, respectively, and many cracks were also observed on the outflow aspect of the disk removed 40 years after implantation. Disk fracture and localized disk wear were found in the SE mitral disk valves implanted >36 years previously. The present results suggest that SE mitral caged-disk valves implanted >20 years previously should be carefully followed up, and that those implanted >30 years previously should be electively replaced with modern prosthetic valves

  1. Maximizing prosthetic valve size with the Top Hat supra-annular aortic valve

    DEFF Research Database (Denmark)

    Aagaard, Jan; Geha, Alexander S.

    2007-01-01

    BACKGROUND AND AIM OF THE STUDY: The CarboMedics Top Hat supra-annular aortic valve allows a one-size (and often two-size) increase over the standard intra-annular valve. This advantage should minimize the risk of patient-prosthesis mismatch, where the effective prosthetic valve orifice area....... This study evaluates the authors' clinical experience with Top Hat supra-annular aortic valve size selection, and the technical aspects of implantation. METHODS: Between January 1999 and October 2005, a total of 251 consecutive patients underwent 252 aortic valve replacements with Top Hat supra...... required unplanned coronary bypass, and 30-day mortality was 2.0% (5/251), indicating a good safety profile for the valves implanted in this series. CONCLUSION: The general distribution of implant sizes in the US indicates that cardiac surgeons may be under-sizing the Top Hat supra-annular aortic valve...

  2. Novel thoughts on patient-prosthesis mismatch in aortic valve replacement: the rationale for the PAR I trial.

    Science.gov (United States)

    Doenst, Torsten; Amorim, Paulo A; Diab, Mahmoud; Hagendorff, Andreas; Faerber, Gloria; Graff, Jürgen; Rastan, Ardawan; Deutsch, Oliver; Eichinger, Walter

    2014-09-01

    The hemodynamic performance of prosthetic tissue valves is influenced by valve design and valve-specific sizing strategies. Design determines the actual geometric opening area (GOA) of the prosthetic valve and sizing strategy its actual chosen size. Currently, hemodynamic performance is assessed by determining the effective orifice area (EOA; derived from the continuity equation by relating flow velocities with the area of the left ventricular outflow tract [LVOTA]). The question whether a valve is too small (patient-prosthesis mismatch [PPM]) is currently addressed by relating EOA to body surface area (EOA index [EOAi]). However, this relation may not be appropriate because the EOAi relates flow velocity to patient-specific anatomic parameters twice (i.e., LVOTA and body surface area). This potential confounder may explain the controversies regarding PPM. However, intuitively, leaving a gradient behind after aortic valve replacement cannot be irrelevant. PPM becomes even more relevant with transcatheter valve-in-valve implantation, where a second prosthesis is taking up inner space of a valve that may have already been too small initially. Thus, a reliable method to determine the presence of PPM is needed. The Prosthesis-to-Annulus Relation I (PAR I) trial is a German multicenter study assessing the relation between the prosthetic GOA and the LVOTA as a potentially new parameter for the prediction of hemodynamic outcome. The results may possibly guide future valve size selection and may allow prediction of functionally relevant PPM. Here, we will demonstrate the shortcomings of the currently applied EOAi for the assessment of hemodynamic relevance and present the rationale for the PARI trial, which recently started recruiting patients. Georg Thieme Verlag KG Stuttgart · New York.

  3. Floppy mitral valve (FMV)/mitral valve prolapse (MVP) and the FMV/MVP syndrome: pathophysiologic mechanisms and pathogenesis of symptoms.

    Science.gov (United States)

    Boudoulas, Konstantinos Dean; Boudoulas, Harisios

    2013-01-01

    Mitral valve prolapse (MVP) results from the systolic movement of a portion or segments of the mitral valve leaflets into the left atrium during left ventricular systole. It is well appreciated today that floppy mitral valve (FMV) is the central issue in the MVP and mitral valve regurgitation (MVR) story. The term FMV refers to the expansion of the area of the mitral valve leaflets with elongated chordae tendineae, chordae rupture and mitral annular dilation. FMV/MVP occurs in a heterogeneous group of patients with a wide spectrum of mitral valve involvement from mild to severe. Two types of symptoms can be defined in FMV/MVP patients. In one group of patients, symptoms are directly related to progressive MVR. In the other group, symptoms cannot be explained by the degree of MVR alone; activation of the autonomic nervous system has been implicated for the explanation of symptoms in this group of patients which is referred to as the FMV/MVP syndrome. In this brief review, the natural history, pathophysiologic mechanisms and management of patients with FMV/MVP/MVR and FMV/MVP syndrome are discussed. © 2013 S. Karger AG, Basel.

  4. Anatomical challenges for transcatheter mitral valve intervention

    DEFF Research Database (Denmark)

    De Backer, Ole; Luk, Ngai H V; Søndergaard, Lars

    2016-01-01

    Following the success of transcatheter aortic and pulmonary valve implantation, there is a large interest in transcatheter mitral valve interventions to treat severe mitral regurgitation (MR). With the exception for the MitraClipTM (Abbott, Abbott Park, IL, USA) edge-to-edge leaflet plication...

  5. Transcatheter Aortic Valve Replacement in Europe

    DEFF Research Database (Denmark)

    Mylotte, Darren; Osnabrugge, Ruben L J; Windecker, Stephan

    2013-01-01

    The authors sought to examine the adoption of transcatheter aortic valve replacement (TAVR) in Western Europe and investigate factors that may influence the heterogeneous use of this therapy.......The authors sought to examine the adoption of transcatheter aortic valve replacement (TAVR) in Western Europe and investigate factors that may influence the heterogeneous use of this therapy....

  6. Oil pipeline valve automation for spill reduction

    Energy Technology Data Exchange (ETDEWEB)

    Mohitpour, Mo; Trefanenko, Bill [Enbridge Technology Inc, Calgary (Canada); Tolmasquim, Sueli Tiomno; Kossatz, Helmut [TRANSPETRO - PETROBRAS Transporte S.A., Rio de Janeiro, RJ (Brazil)

    2003-07-01

    Liquid pipeline codes generally stipulate placement of block valves along liquid transmission pipelines such as on each side of major river crossings where environmental hazards could cause or are foreseen to potentially cause serious consequences. Codes, however, do not stipulate any requirement for block valve spacing for low vapour pressure petroleum transportation, nor for remote pipeline valve operations to reduce spills. A review of pipeline codes for valve requirement and spill limitation in high consequence areas is thus presented along with a criteria for an acceptable spill volume that could be caused by pipeline leak/full rupture. A technique for deciding economically and technically effective pipeline block valve automation for remote operation to reduce oil spilled and control of hazards is also provided. In this review, industry practice is highlighted and application of the criteria for maximum permissible oil spill and the technique for deciding valve automation thus developed, as applied to ORSUB pipeline is presented. ORSUB is one of the three initially selected pipelines that have been studied. These pipelines represent about 14% of the total length of petroleum transmission lines operated by PETROBRAS Transporte S.A. (TRANSPETRO) in Brazil. Based on the implementation of valve motorization on these three pipeline, motorization of block valves for remote operation on the remaining pipelines is intended, depending on the success of these implementations, on historical records of failure and appropriate ranking. (author)

  7. Nuclear valves and the licencing authority

    International Nuclear Information System (INIS)

    Stevens-Guille, P.D.

    1977-02-01

    Valve problems relevant to licencing authorities are discussed. It is suggested that the probability of valve failure should be given more emphasis in safety analysis. Problems of stress analysis, codes, seismic effects, malfunction and leakage are discussed and programmes aimed at solving future problems are outlined

  8. The future of transcatheter mitral valve interventions

    DEFF Research Database (Denmark)

    Maisano, Francesco; Alfieri, Ottavio; Banai, Shmuel

    2015-01-01

    of transcatheter mitral valve interventions will be. The purpose of the present report is to review the current state-of-the-art of mitral valve intervention, and to identify the potential future scenarios, which might benefit most from the transcatheter repair and replacement devices under development....

  9. Mitral valve prolapse - report of 3 cases

    International Nuclear Information System (INIS)

    Han, Moon Hee; Im, Chung Ki; Im, Dong Ran; Han, Man Chung; Lee, Young Woo; Seo, Jung Don

    1979-01-01

    Prolapse of mitral valve is characterized by its unique auscultatory, echocardiographic and angiographic findings and may be associated with various disease entities such as congenital heart disease, coronary heart disease and Marfan's syndrome etc. Authors report recent experience of 3 cases of prolapsed mitral valve, 2 cases associated with A.S.D. and 1 case with Marfan's syndrome.

  10. Altitude valve for railway suspension control system

    Science.gov (United States)

    Zhang, Xuan; Zhang, Lihao; Li, Qingxuan; Chen, WanSong

    2017-09-01

    With the variation of people and material during vehicle service, the gravity of vehicle could be unbalanced. As a result it might cause accident. In order to solve this problem, altitude valve is assembled on board. It can adjust the gravity of vehicle by the intake and outlet progress of the spring in the altitude valve to prevent the tilt of vehicles.

  11. Bentall Operation with Valved Homograft Conduit

    Science.gov (United States)

    Choudhary, Shiv K.; Talwar, Sachin; Kumar, A. Sampath

    2000-01-01

    Lesions of the ascending aorta associated with aortic valve disease are usually treated by implanting a prosthetic valved conduit (Bentall procedure). In this report, we present our experience in which a valved homograft conduit was used for the procedure. Six patients underwent a Bentall procedure with the use of a cryopreserved valved homograft conduit. Two of the patients had annuloaortic ectasia, 2 had Marfan syndrome, and 1 had an atherosclerotic aneurysm of the aorta. One patient had severe aortic stenosis due to a bicuspid aortic valve, along with an aneurysm and localized dissection of the ascending aorta. In all of the patients, the aortic annulus was substantially dilated, with accompanying moderate-to-severe aortic regurgitation. A standard procedure was performed with moderate hypothermia, cardiopulmonary bypass, and aortic and bicaval cannulation. The ascending aorta and the aortic valve were replaced with a cryopreserved valved homograft conduit (aortic in 5 patients and pulmonary in 1). The native coronary ostia were anastomosed directly to the homograft. Echocardiography, which was performed intraoperatively, before discharge from the hospital, and at follow-up visits (1 to 36 months), revealed good valve function without dilatation of the homograft conduits. There was 1 late death due to Aspergillus fumigatus endocarditis, 6 months postoperatively. In 1 patient, magnetic resonance imaging performed at 24 months revealed normal caliber of the homograft conduit. We conclude that the Bentall procedure can be performed, safely and with excellent results, using cryopreserved homograft conduits. PMID:11198310

  12. 14 CFR 25.995 - Fuel valves.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 1 2010-01-01 2010-01-01 false Fuel valves. 25.995 Section 25.995 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF TRANSPORTATION AIRCRAFT AIRWORTHINESS STANDARDS: TRANSPORT CATEGORY AIRPLANES Powerplant Fuel System Components § 25.995 Fuel valves. In addition...

  13. 14 CFR 29.995 - Fuel valves.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 1 2010-01-01 2010-01-01 false Fuel valves. 29.995 Section 29.995 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF TRANSPORTATION AIRCRAFT AIRWORTHINESS STANDARDS: TRANSPORT CATEGORY ROTORCRAFT Powerplant Fuel System Components § 29.995 Fuel valves. In addition...

  14. 14 CFR 27.995 - Fuel valves.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 1 2010-01-01 2010-01-01 false Fuel valves. 27.995 Section 27.995 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT OF TRANSPORTATION AIRCRAFT AIRWORTHINESS STANDARDS: NORMAL CATEGORY ROTORCRAFT Powerplant Fuel System Components § 27.995 Fuel valves. (a) There must...

  15. Heart valve disease among patients with hyperprolactinemia

    DEFF Research Database (Denmark)

    Steffensen, Maria Charlotte; Maegbaek, Merete Lund; Laurberg, Peter

    2012-01-01

    Increased risk of heart valve disease during treatment with certain dopamine agonists, such as cabergoline, has been observed in patients with Parkinson's disease. The same compound is used to treat hyperprolactinemia, but it is unknown whether this also associates with heart valve disease....

  16. Evolving Concepts in Transcatheter Aortic Valve Implantation

    NARCIS (Netherlands)

    Nijhoff, F.

    2015-01-01

    Part I of the present thesis is dedicated to implantation technique and the clinical performance of new valve prostheses. A satisfactory TAVI result not solely depends on patient characteristics, but also relies on proper valve positioning and final placement. Moreover, prosthetic design is

  17. Hardfacing materials and processes for valve applications

    International Nuclear Information System (INIS)

    Matthews, S.J.; Crook, P.

    1982-01-01

    The subject of hardfacing is a very high technology effort especially in the valve industry. The technology is manifested by the need for sophisticated high performance hardfacing alloys required to resist the demanding environments of fluid flow control valve service. High technology is also found in the automated methods currently being used to efficiently deposit high quality hardfacing overlays. 3 figures, 3 tables

  18. Heart valve replacements with regenerative capacity

    NARCIS (Netherlands)

    Dijkman, P.E.; Fioretta, E.S.; Frese, L.; Pasqualini, F.S.; Hoerstrup, S.P.

    2016-01-01

    The incidence of severe valvular dysfunctions (e.g., stenosis and insufficiency) is increasing, leading to over 300,000 valves implanted worldwide yearly. Clinically used heart valve replacements lack the capacity to grow, inherently requiring repetitive and high-risk surgical interventions during

  19. Butterfly valves: greater use in power plants

    International Nuclear Information System (INIS)

    McCoy, M.

    1975-01-01

    Improvements in butterfly valves, particularly in the areas of automatic control and leak tightness are described. The use of butterfly valves in nuclear power plants is discussed. These uses include service in component cooling, containment cooling, and containment isolation. The outlook for further improvements and greater uses is examined. (U.S.)

  20. Controllable valve in a nuclear reactor system

    International Nuclear Information System (INIS)

    Schabert, H.P.; Laurer, E.

    1980-01-01

    The quick-acting gate valve of the PWR is opened and closed by means of two pistons and live steam. One of the pistons is connected to the valve disk by a piston rod which is concentrically lead into another hollow piston rod being connected to the second piston. Stops limit the strokes of the two pistons. (GL) [de

  1. How to insure quality valve remanufacture

    International Nuclear Information System (INIS)

    Scott, C.F.

    1991-01-01

    The importance of quality valve repair for the power generation industry is an obvious need for both the owner as well as the consumer. Whether valves are repaired in-line, on-site, or at a valve remanufacturing facility, the selection of a vendor is vital to meeting not only stringent quality requirements, but also to meet start-up schedules and budgets. In the past, the rule of thumb was that repair of a valve could cost approximately 50% of the cost of a new valve and still represent a significant savings to the end user. For power generation facilities, the fact that many valves are welded in not only makes repair more economical, but even vital to continuing normal operations. For those items not welded in, long lead times and higher prices for these normally exotic alloys make remanufactured valves even more attractive. However, even as these advantages of remanufacturing are obvious, some repair organizations continue to cut corners to meet profit demands. The result is suspect quality in some valves. This can lead to premature failures, possible reduced generating capacity, unscheduled outages, and even catastrophic results. Therefore, the choice of a repair organization must be made with care. As the author has said, repair is an obvious option, but the procurement should definitely involve more than just price comparisons. Evaluation must place the emphasis on quality and reliability. Several aspects should be thoroughly investigated and documented in the selection process. These include: personnel; equipment/facilities; procedures; and credentials

  2. Vertical pump with free floating check valve

    International Nuclear Information System (INIS)

    Lindsay, M.

    1980-01-01

    A vertical pump is described which has a bottom discharge with a free floating check valve disposed in the outlet plenum thereof. The free floating check valve comprises a spherical member with a hemispherical cage-like member attached thereto which is capable of allowing forward or reverse flow under appropriate conditions while preventing reverse flow under inappropriate conditions

  3. Experience with the Ahmed Glaucoma Valve

    African Journals Online (AJOL)

    Design:Aprospective study of three glaucoma patients who had the Ahmed glaucoma valve implant at the. University of Benin ... Key words: glaucoma, glaucoma valve implant, ... introduced the micro-trephination with a diameter of. 0.6mm in ...

  4. Bentall procedure using cryopreserved valved aortic homografts: mid- to long-term results.

    Science.gov (United States)

    Christenson, Jan T; Sierra, Jorge; Trindade, Pedro T; Dominique, Didier; Kalangos, Afksendiyos

    2004-01-01

    The Bentall procedure is the standard operation for patients who have lesions of the ascending aorta associated with aortic valve disease. In many cases, however, mechanical prosthetic conduits are not suitable. There are few reports in the English-language medical literature concerning the mid- to long-term outcome of Bentall operations with cryopreserved homografts. Therefore, we reviewed our experience with this procedure and valved homografts. From January 1997 through December 2002, 21 patients underwent a Bentall operation with cryopreserved homografts at our institution. There were 14 males and 7 females; the mean age was 36 +/- 21 years (range, 15-74 years). Eleven patients had undergone previous aortic valve surgery. All patients had aortic dilatation or aneurysms involving the ascending aorta. Indications for surgery included aortic valve stenosis or insufficiency, and aortic valve endocarditis (native valve or prosthetic). One patient had Takayasu's arteritis and 3 had Marfan syndrome. There was 1 hospital death (due to sepsis), but no other major postoperative complications. The mean hospital stay was 14 +/- 7 days. Follow-up echocardiographic and computed tomographic scans were performed yearly. The mean follow-up was 34 months (6-72 months). Follow-up imaging revealed no calcifications or degenerative processes related to the homograft. Four patients had minimal valve regurgitation. Two patients died during follow-up. The 3-year actuarial survival rate was 85.7%. Our data suggest that the Bentall procedure with a valved homograft conduit is a safe procedure with excellent mid- to long-term results, comparable to results reported with aortic valve replacement with a homograft.

  5. Defining indicators to motorize block valves aiming to reduce potential leakage applied to OSBRA pipeline

    Energy Technology Data Exchange (ETDEWEB)

    Pires, L.F.G. [Pontificia Universidade Catolica do Rio de Janeiro (PUC-Rio), RJ (Brazil); Sousa, A.G.; Castro, N.C.; Spagnolo, R. [Petrobras Transporte S.A. (TRANSPETRO), Rio de Janeiro, RJ (Brazil)

    2009-07-01

    The discussion about motorizing block valves is a constant point being brought up when it is intended to control or reduce the amount of liquid leaking in the event of a pipe rupture. During the pipeline's project stage the installation of blocking valves along the pipeline must be taken into consideration to meet the operation and maintenance requirements as well as to reduce the potential amounts of volume being leaked. In existing pipelines, the main concern is the definition of which valves are candidates to be motorized. In both situations criteria should be established to define this choice. A math algorithm was developed to define the potential leakage due to gravity along the pipeline profile where the influence of a valve over another is verified, as well as the contribution of the check valves existing in the pipeline. The present work defines a parameter based on the extension protected by the valve and the reduction of the potential leakage. This parameter is then fed to a worksheet where the efficiency indicators are calculated to each valve eligible to be motorized. It also takes into consideration factors relative to the valve location, such as the environmental sensitivity, risk assessment, social diagnosis and device's proximity to contingency resources. Finally, after considering all the above aspects, it's possible to come up with a final classification, recommending specific valves to be prioritized on an eventual process of motorization adequacies. This methodology was applied, experimentally on a pipe segment of TRANSPETRO's Sao Paulo-Brasilia pipeline - OSBRA , where it proved to be an important technological and management tool. (author)

  6. Leaving Moderate Tricuspid Valve Regurgitation Alone at the Time of Pulmonary Valve Replacement: A Worthwhile Approach.

    Science.gov (United States)

    Kogon, Brian; Mori, Makoto; Alsoufi, Bahaaldin; Kanter, Kirk; Oster, Matt

    2015-06-01

    Pulmonary valve disruption in patients with tetralogy of Fallot and congenital pulmonary stenosis often results in pulmonary insufficiency, right ventricular dilation, and tricuspid valve regurgitation. Management of functional tricuspid regurgitation at the time of subsequent pulmonary valve replacement remains controversial. Our aims were to (1) analyze tricuspid valve function after pulmonary valve replacement through midterm follow-up and (2) determine the benefits, if any, of concomitant tricuspid annuloplasty. Thirty-five patients with tetralogy of Fallot or congenital pulmonary stenosis were analyzed. All patients had been palliated in childhood by disrupting the pulmonary valve, and all patients had at least moderate tricuspid valve regurgitation at the time of subsequent pulmonary valve replacement. Preoperative and serial postoperative echocardiograms were analyzed. Pulmonary and tricuspid regurgitation, along with right ventricular dilation and dysfunction were scored as 0 (none), 1 (mild), 2 (moderate), and 3 (severe). Right ventricular volume and area were also calculated. Comparisons were made between patients who underwent pulmonary valve replacement alone and those who underwent concomitant tricuspid valve annuloplasty. At 1 month after pulmonary valve replacement, there were significant reductions in pulmonary valve regurgitation (mean 3 vs 0.39, p tricuspid valve regurgitation (mean 2.33 vs 1.3, p tricuspid regurgitation 1 month postoperatively between patients who underwent concomitant tricuspid annuloplasty and those who underwent pulmonary valve replacement alone (mean 1.31 vs 1.29, p = 0.81). However, at latest follow-up (mean 7.0 ± 2.8 years), the degree of tricuspid regurgitation was significantly higher in the concomitant annuloplasty group (mean 1.87 vs 1.12, p = 0.005). In patients with at least moderate tricuspid valve regurgitation, significant improvement in tricuspid valve function and right ventricular size occurs in the first

  7. Aortic Root Enlargement or Sutureless Valve Implantation?

    Directory of Open Access Journals (Sweden)

    Nikolaos G. Baikoussis

    2016-11-01

    Full Text Available Aortic valve replacement (AVR in patients with a small aortic annulus is a challenging issue. The importance of prosthesis–patient mismatch (PPM post aortic valve replacement (AVR is controversial but has to be avoided. Many studies support the fact that PPM has a negative impact on short and long term survival. In order to avoid PPM, aortic root enlargement may be performed. Alternatively and keeping in mind that often some comorbidities are present in old patients with small aortic root, the Perceval S suturelles valve implantation could be a perfect solution. The Perceval sutureless bioprosthesis provides reasonable hemodynamic performance avoiding the PPM and providing the maximum of aortic orifice area. We would like to see in the near future the role of the aortic root enlargement techniques in the era of surgical implantation of the sutureless valve (SAVR and the transcatheter valve implantation (TAVI.

  8. Transcatheter Mitral Valve Devices - Functional Mechanical Designs.

    Science.gov (United States)

    Kliger, Chad

    2014-03-01

    Mitral regurgitation is a complex disorder involving a multitude of components of the mitral apparatus. With the desire for less invasive treatment approaches, transcatheter mitral valve therapies (TMVT) are directed at these components and available at varying stages of development. Therapeutic advancements and the potential to combine technologies may further improve their efficacy and safety. Transcatheter mitral valve replacement, while preserving the mitral apparatus, may emerge as an alternative or even a more suitable treatment option. In addition, early data on transcatheter mitral valve-in-valve and valve-in-ring implantation are encouraging and this approach may be an alternative to reoperation in the high-risk patient. This review details the expanding functional mechanical designs of current active TMVT.

  9. Biological aortic valve replacement: advantages and optimal indications of stentless compared to stented valve substitutes. A review.

    Science.gov (United States)

    Tavakoli, Reza; Danial, Pichoy; Oudjana, Ahmed Hamid; Jamshidi, Peiman; Gassmann, Max; Leprince, Pascal; Lebreton, Guillaume

    2018-05-01

    Controversy still surrounds the optimal biological valve substitute for aortic valve replacement. In light of the current literature, we review advantages and optimal indications of stentless compared to stented aortic bio-prostheses. Recent meta-analyses, prospective randomized controlled trials and retrospective studies comparing the most frequently used stentless and stented aortic bio-prostheses were analyzed. In the present review, the types and implantation techniques of the bio-prosthesis that are seldom taken into account by most studies and reviews were integrated in the interpretation of the relevant reports. For stentless aortic root bio-prostheses, full-root vs. sub-coronary implantation offered better early transvalvular gradients, effective orifice area and left ventricular mass regression as well as late freedom from structural valve deterioration in retrospective studies. Early mortality and morbidity did not differ between the stentless and stented aortic bio-prostheses. Early transvalvular gradients, effective orifice area and regression of left ventricular hypertrophy were significantly better for stentless, especially as full-root, compared to stented bio-prostheses. The long-term valve-related survival for stentless aortic root and Toronto SPV bio-prosthesis was as good as that for stented pericardial aortic bio-prostheses. For full-root configuration this survival advantage was statistically significant. There seems to be not one but different ideal biological valve substitutes for different subgroups of patients. In patients with small aortic root or exposed to prosthesis-patient mismatch full-root implantation of stentless bio-prostheses may better meet functional needs of individual patients. Longer follow-ups on newer generation of stented bio-prostheses are needed for comparison of their hemodynamic performance with stentless counterparts especially in full-root configuration.

  10. Association of aortic valve calcification severity with the degree of aortic regurgitation after transcatheter aortic valve implantation.

    Science.gov (United States)

    Koos, Ralf; Mahnken, Andreas Horst; Dohmen, Guido; Brehmer, Kathrin; Günther, Rolf W; Autschbach, Rüdiger; Marx, Nikolaus; Hoffmann, Rainer

    2011-07-15

    This study sought to examine a possible relationship between the severity of aortic valve calcification (AVC), the distribution of AVC and the degree of aortic valve regurgitation (AR) after transcatheter aortic valve implantation (TAVI) for severe aortic stenosis (AS). 57 patients (22 men, 81 ± 5 years) with symptomatic AS and with a logistic EuroSCORE of 24 ± 12 were included. 38 patients (67%) received a third (18F)-generation CoreValve® aortic valve prosthesis, in 19 patients (33%) an Edwards SAPIEN™ prosthesis was implanted. Prior to TAVI dual-source computed tomography for assessment of AVC was performed. To determine the distribution of AVC the percentage of the calcium load of the most severely calcified cusp was calculated. After TAVI the degree of AR was determined by angiography and echocardiography. The severity of AR after TAVI was related to the severity and distribution of AVC. There was no association between the distribution of AVC and the degree of paravalvular AR after TAVI as assessed by angiography (r = -0.02, p = 0.88). Agatston AVC scores were significantly higher in patients with AR grade ≥ 3 (5055 ± 1753, n = 3) than in patients with AR grade AVC scores > 3000 were associated with a relevant paravalvular AR and showed a trend for increased need for second manoeuvres. There was a significant correlation between the severity of AVC and the degree of AR after AVR (r = 0.50, p AVC have an increased risk for a relevant AR after TAVI as well as a trend for increased need for additional procedures. Copyright © 2010 Elsevier Ireland Ltd. All rights reserved.

  11. Vortex dynamics in Patient-Specific Stenotic Tricuspid and Bicuspid Aortic Valves pre- and post- Trans-catheter Aortic Valve Replacement

    Science.gov (United States)

    Hatoum, Hoda; Dasi, Lakshmi Prasad

    2017-11-01

    Understanding blood flow related adverse complications such as leaflet thrombosis post-transcatheter aortic valve implantation (TAVI) requires a deeper understanding of how patient-specific anatomic and hemodynamic factors, and relative valve positioning dictate sinus vortex flow and stasis regions. High resolution time-resolved particle image velocimetry measurements were conducted in compliant and transparent 3D printed patient-specific models of stenotic bicuspid and tricuspid aortic valve roots from patients who underwent TAVI. Using Lagrangian particle tracking analysis of sinus vortex flows and probability distributions of residence time and blood damage indices we show that (a) patient specific modeling provides a more realistic assessment of TAVI flows, (b) TAVI deployment alters sinus flow patterns by significantly decreasing sinus velocity and vorticity, and (c) relative valve positioning can control critical vortex structures that may explain preferential leaflet thrombosis corresponding to separated flow recirculation, secondary to valve jet vectoring relative to the aorta axis. This work provides new methods and understanding of the spatio-temporal aortic sinus vortex dynamics in post TAVI pathology. This study was supported by the Ohio State University DHLRI Trifit Challenge award.

  12. Aging and malfunction of valves in CANDU special safety systems. Phase 1

    International Nuclear Information System (INIS)

    1989-03-01

    Aging and wear related valve malfunctions have been reported in American nuclear generating systems. This report documents the first attempt to study these phenomena on a global basis in Canadian nuclear power plants. A general methodology outlines an approach to this type of study which is amenable to use within existing information structures. Nuclear regulatory requirements which influence the testing of valves in Canadian nuclear power plants are reviewed. The reporting systems which emanate from these requirements are discussed and sources of valve failure data are reviewed. It is determined that modifications to existing failure reporting systems are required before practical means of collecting data necessary for the analysis of age related valve malfunctions can be developed. In spite of limitations in reported failure data, a partial data base is compiled for valve failures in Special Safety Systems of domestic nuclear plants. Data are reported for the period 1982 to 1986. The valve population and basic parameters of each valve such as type, operator, function, etc., and the reported failures against this population are compiled and reviewed for evidence of time dependent versus random failure trends. Results suggest that there is no clear age related failure trend. In fact, some systems and stations, experienced a reduction in failure rates with years of servicing, suggesting that some earlier generic valve problems may have been solved. Present inspection, test, and maintenance practices are reviewed and their effectiveness for purposes of predicting or preventing incipient failures is assessed to be of moderate value. Modern failure prevention methods are highlighted and their applicability discussed

  13. Risk evaluation for motor operated valves in an Inservice Testing Program at a PWR nuclear power plant in Taiwan

    International Nuclear Information System (INIS)

    Li, Y.C.; Chen, K.T.; Su, Y.L.; Ting, K.; Chien, F.T.; Li, G.D.; Huang, S.H.

    2012-01-01

    Safety related valves such as Motor Operated Valves (MOV), Air Operated Valves (AOV) or Check Valves (CV) play an important role in nuclear power plant. Functioning of these valves mainly aim at emergency reactivity control, post-accident residue heat removal, post-accident radioactivity removal and containment isolation when a design basis accident occurred. In order to maintain these valves under operable conditions, an Inservice Testing Program (IST) is defined for routine testing tasks based on the ASME Boiler and Pressure Vessel Code section XI code requirements. Risk based Inservice Testing Programs have been studied and developed extensively in the nuclear energy industry since the 1990s. Risk Based evaluations of IST can bring positive advantages to the licensee such as identifying the vulnerability of the system, reducing unnecessary testing burden, concentrating testing resources on the critical pass oriented valves and saving plant’s personnel dose exposure. This risk evaluation is incorporated with quantitative and qualitative analyses to the Motor Operated Valves under current Inservice Testing Program for PWR nuclear power plant in Taiwan. With the outcome of the risk classifications for the safety related MOVs through numerical or deterministic analyses, a risk based testing frequency relief is suggested to demonstrate the benefits received from the risk based Inservice Testing Program. The goal made of this study, it could be as a reference and cornerstone for the licensee to perform overall scope Risk-Informed Inservice Testing Program (RI-IST) evaluation by referring relevant methodologies established in this study.

  14. Use of a valve operation test and evaluation system to enhance valve reliability

    International Nuclear Information System (INIS)

    Lowry, D.A.

    1990-01-01

    Power plant owners have emphasized the need for assuring safe, reliable operation of valves. While most valves must simply open or close, the mechanisms involved can be quite complex. Motor operated valves (MOVs) must be properly adjusted to assure operability. Individual operator components determine the performance of the entire MOV. Failure in MOVs could cripple or shut down a unit. Thus, a complete valve program consisting of design reviews, operational testing, and preventive and predictive maintenance activities will enhance an owner's confidence level that his valves win operate as expected. Liberty's Valve Operation Test and Evaluation System (VOTES) accurately measures stein thrust without intruding on valve operation. Since mounting a strain gage to a valve stem is a desirable but impractical way of obtaining precise stem thrust, Liberty developed a method to obtain identical data by placing a strain gage sensor on the valve yoke. VOTES provides information which effectively eliminates costly, unscheduled downtime. This paper presents the results of infield VOTES testing. The system's proven ability to identify and characterize actuator and valve performance is demonstrated. Specific topics of discussion include the ability of VOTES to ease a utility's IE Bulletin 8543 concerns and conclusively diagnose MOV components. Data from static and differential pressure testing are presented. Technical, operational, and financial advantages resulting from VOTES technology are explored in detail

  15. THE RESULTS OF SURGICAL TREATMENT OF TRICUSPID VALVE INFECTIVE ENDOCARDITIS USING VALVE REPAIR AND VALVE REPLACEMENT OPERATIONS

    Directory of Open Access Journals (Sweden)

    S. A. Kovalev

    2015-01-01

    Full Text Available Aim. To evaluate in-hospital and long-term results of surgical treatment of patients with infective endocarditis of the tricuspid valve, to compare the effectiveness of valve repair and valve replacement techniques, and to identify risk factors of mortality and reoperations. Materials and methods. 31 surgical patients with tricuspid valve infective endocarditis were evaluated. Patients were divided into 2 groups. In Group 1 (n = 14 repairs of the tricuspid valve were performed, in Group 2 (n = 17 patients had undergone tricuspid valve replacements. Epidemiological, clinical, microbiological and echocardiographic data were studied. Methods of comparative analysis, the Kaplan–Meier method, and Cox risk models were applied. Results. The most common complication of in-hospital stay was atrioventricular block (17.7% of cases in Group 2. In Group 1, this type of complication was not found. Hospital mortality was 7.14% in Group 1, and 0% in Group 2. Long-term results have shown the significant reduction of heart failure in general cohort and in both groups. In Group 1 the severity of heart failure in the long term was less than in Group 2. No significant differences in the severity of tricuspid regurgitation were found between the groups. In 7-year follow up no cases of death were registered in Group 1. Cumulative survival rate in Group 2 within 60 months was 67.3 ± 16.2%. No reoperations were performed in patients from Group 1. In Group 2, the freedom from reoperation within 60 months was 70.9 ± 15.3%. Combined intervention was found as predictor of postoperative mortality. Prosthetic valve endocarditis was identified as risk factor for reoperation. Conclusion. Valve repair and valve replacement techniques of surgical treatment of tricuspid valve endocarditis can provide satisfactory hospital and long-term results. Tricuspid valve repair techniques allowed reducing the incidence of postoperative atrioventricular block. In the long-term, patients

  16. A method for evaluating pressure locking and thermal binding of gate valves

    Energy Technology Data Exchange (ETDEWEB)

    Dogan, T.

    1996-12-01

    A method is described to evaluate the susceptibility of gate valves to pressure locking and thermal binding. Binding of the valve disc in the closed position due to high pressure water trapped in the bonnet cavity (pressure locking) or differential thermal expansion of the disk in the seat (thermal binding) represents a potential mechanism that can prevent safety-related systems from functioning when called upon. The method described here provides a general equation that can be applied to a given gate valve design and set of operating conditions to determine the susceptibility of the valve to fail due to disc binding. The paper is organized into three parts. The first part discusses the physical mechanisms that cause disc binding. The second part describes the mathematical equations. The third part discusses the conclusions.

  17. [Valve-sparing aortic root replacement for young female patients with Marfan syndrome].

    Science.gov (United States)

    Nawata, Kan; Morota, T

    2009-10-01

    Annuloaortic ectasia is frequently related with Marfan syndrome, and Bentall procedure or aortic root replacement with valved conduit has been the conventional standard operation for this disease. Recently, some institutes have adopted valve-sparing aortic root replacement (VSARR) instead of Bentall procedure. Young female patients with Marfan syndrome who wishes for childbearing seem to be a group of good candidates of this type of operation, because it let them free from morbidities after artificial valve implantation. Valve-sparing operation should be taken into consideration when the size of Valsalva sinus reaches 45 mm for patients with Marfan syndrome and when it reaches 40 mm for patients with past histories or family histories of aortic dissection or aortic rupture. Since pregnancy is one of the most serious risk factors for aortic events, Valsalva sinus of 40 mm or larger could be the new standard for surgical indication if VSARR is applicable.

  18. Design, fabrication and characterization of an arrayable all-polymer microfluidic valve employing highly magnetic rare-earth composite polymer

    International Nuclear Information System (INIS)

    Rahbar, Mona; Gray, Bonnie L; Shannon, Lesley

    2016-01-01

    We present a new magnetically actuated microfluidic valve that employs a highly magnetic composite polymer (M-CP) containing rare-earth hard-magnetic powder for its actuating element and for its valve seat. The M-CP offers much higher magnetization compared to the soft-magnetic, ferrite-based composite polymers typically used in microfluidic applications. Each valve consists of a permanently magnetized M-CP flap and valve seat mounted on a microfluidic channel system fabricated in poly(dimethylsiloxane) (PDMS). Each valve is actuated under a relatively small external magnetic field of 80 mT provided by a small permanent magnet mounted on a miniature linear actuator. The performance of the valve with different flap thicknesses is characterized. In addition, the effect of the magnetic valve seat on the valve’s performance is also characterized. It is experimentally shown that a valve with a 2.3 mm flap thickness, actuated under an 80 mT magnetic field, is capable of completely blocking liquid flow at a flow rate of 1 ml min −1 for pressures up to 9.65 kPa in microfluidic channels 200 μ m wide and 200 μ m deep. The valve can also be fabricated into an array for flow switching between multiple microfluidic channels under continuous flow conditions. The performance of arrays of valves for flow routing is demonstrated for flow rates up to 5 ml min −1 with larger microfluidic channels of up to 1 mm wide and 500 μ m deep. The design of the valves is compatible with other commonly used polymeric microfluidic components, as well as other components that use the same novel permanently magnetic composite polymer, such as our previously reported cilia-based mixing devices. (paper)

  19. Environmental qualification testing of TFE valve components

    International Nuclear Information System (INIS)

    Eyvindson, A.; Krasinski, W.; McCutcheon, R.

    1997-01-01

    Valves containing tetrafluoroethylene (TFE) components are being used in many CANDU Nuclear Generating Stations. However, some concerns remain about the performance of TFE after exposure to high levels of radiation. Stations must therefore ensure that such valves perform reliably after being exposed to postulated accident radiation dose levels. The current Ontario Hydro Environmental Qualification [EQ] program specifies much higher postulated radiation exposure than the original design, to account for conditions following a LOCA. Initial assessments indicated that Teflon components would require replacement. Proof of acceptable performance can remove the need for large scale replacement, avoiding a significant cost penalty and preserving benefits due to the superior performance of TFE-based seals. A test program was undertaken at Chalk River Laboratories (CRL) to investigate the performance of three valves after irradiation to 10 Mrad. Such valves are currently used at the Bruce B Nuclear Generating Station. Each contains TFE packing rings; one also has TFE seats. Two of the valves are used in the ECIS recovery system, while the third is used for instrumentation loop isolation or as drain valves. All are exposed to little or no radiation during normal use. Based on the results of the tests, all the valves tested will still meet functional and performance requirements after the TFE components have been exposed to 10 Mrad of irradiation. (author)

  20. Motor operated valves problems tests and simulations

    Energy Technology Data Exchange (ETDEWEB)

    Pinier, D.; Haas, J.L.

    1996-12-01

    An analysis of the two refusals of operation of the EAS recirculation shutoff valves enabled two distinct problems to be identified on the motorized valves: the calculation methods for the operating torques of valves in use in the power plants are not conservative enough, which results in the misadjustement of the torque limiters installed on their motorizations, the second problem concerns the pressure locking phenomenon: a number of valves may entrap a pressure exceeding the in-line pressure between the disks, which may cause a jamming of the valve. EDF has made the following approach to settle the first problem: determination of the friction coefficients and the efficiency of the valve and its actuator through general and specific tests and models, definition of a new calculation method. In order to solve the second problem, EDF has made the following operations: identification of the valves whose technology enables the pressure to be entrapped: the tests and numerical simulations carried out in the Research and Development Division confirm the possibility of a {open_quotes}boiler{close_quotes} effect: determination of the necessary modifications: development and testing of anti-boiler effect systems.

  1. Hardfacing and packings for improved valve performance

    International Nuclear Information System (INIS)

    Aikin, J.A.; Patrick, J.N.F.; Inglis, I.

    2003-01-01

    The CANDU Owners Group (COG), Chemistry, Materials and Components (CMC) Program has supported an ongoing program on valve maintenance and performance for several years. An overview is presented of recent work on iron-based hardfacing, packing qualification, friction testing of polytetrafluoroethylene (PTFE) packings, and an investigation of re-torquing valve packing. Based on this program, two new valve-packing materials have been qualified for use in CANDU stations. By doing this, CANDU maintenance can avoid having only one packing qualified for station use, as well as assess the potential impact of the industry trend towards using lower gland loads. The results from corrosion tests by AECL and the coefficient of friction studies at Battelle' s tribology testing facilities on Delcrome 910, an iron-based hardfacing alloy, indicate it is an acceptable replacement for Stellite 6 under certain conditions. This information can be used to update in-line valve purchasing specifications. The renewed interest in friction characteristics, and environmental qualification (EQ) of packing containing PTFE has resulted in a new test program in these areas. The COG-funded valve programs have resulted in modifications to design specifications for nuclear station in-line valves and have led to better maintenance practices and valve reliability. In the end, this means lower costs and cheaper electricity. (author)

  2. Motor operated valves problems tests and simulations

    International Nuclear Information System (INIS)

    Pinier, D.; Haas, J.L.

    1996-01-01

    An analysis of the two refusals of operation of the EAS recirculation shutoff valves enabled two distinct problems to be identified on the motorized valves: the calculation methods for the operating torques of valves in use in the power plants are not conservative enough, which results in the misadjustement of the torque limiters installed on their motorizations, the second problem concerns the pressure locking phenomenon: a number of valves may entrap a pressure exceeding the in-line pressure between the disks, which may cause a jamming of the valve. EDF has made the following approach to settle the first problem: determination of the friction coefficients and the efficiency of the valve and its actuator through general and specific tests and models, definition of a new calculation method. In order to solve the second problem, EDF has made the following operations: identification of the valves whose technology enables the pressure to be entrapped: the tests and numerical simulations carried out in the Research and Development Division confirm the possibility of a open-quotes boilerclose quotes effect: determination of the necessary modifications: development and testing of anti-boiler effect systems

  3. Flowfield Analysis of a Pneumatic Solenoid Valve

    Directory of Open Access Journals (Sweden)

    Sheam-Chyun Lin

    2018-07-01

    Full Text Available Pneumatic solenoid valve has been widely used in the vehicle control systems for meeting the rapid-reaction demand triggered by the dynamic conditions encountered during the driving course of vehicle. For ensuring the safety of human being, the reliable and effective solenoid valve is in great demand to shorten the reaction time and thus becomes the topic of this research. This numerical study chooses a commercial 3/2-way solenoid valve as the reference valve for analysing its performance. At first, CFD software Fluent is adopted to simulate the flow field associated with the valve configuration. Then, the comprehensive flow visualization is implemented to identify the locations of adverse flow patterns. Accordingly, it is found that a high-pressure region exists in the zone between the nozzle exit and the top of iron core. Thereafter, the nozzle diameter and the distance between nozzle and spool are identified as the important design parameters for improving the pressure response characteristics of valve. In conclusion, this work establishes a rigorous and systematic CFD scheme to evaluate the performance of pneumatic solenoid valve.

  4. Automated valve fault detection based on acoustic emission parameters and support vector machine

    Directory of Open Access Journals (Sweden)

    Salah M. Ali

    2018-03-01

    Full Text Available Reciprocating compressors are one of the most used types of compressors with wide applications in industry. The most common failure in reciprocating compressors is always related to the valves. Therefore, a reliable condition monitoring method is required to avoid the unplanned shutdown in this category of machines. Acoustic emission (AE technique is one of the effective recent methods in the field of valve condition monitoring. However, a major challenge is related to the analysis of AE signal which perhaps only depends on the experience and knowledge of technicians. This paper proposes automated fault detection method using support vector machine (SVM and AE parameters in an attempt to reduce human intervention in the process. Experiments were conducted on a single stage reciprocating air compressor by combining healthy and faulty valve conditions to acquire the AE signals. Valve functioning was identified through AE waveform analysis. SVM faults detection model was subsequently devised and validated based on training and testing samples respectively. The results demonstrated automatic valve fault detection model with accuracy exceeding 98%. It is believed that valve faults can be detected efficiently without human intervention by employing the proposed model for a single stage reciprocating compressor. Keywords: Condition monitoring, Faults detection, Signal analysis, Acoustic emission, Support vector machine

  5. 49 CFR 192.193 - Valve installation in plastic pipe.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 3 2010-10-01 2010-10-01 false Valve installation in plastic pipe. 192.193... Components § 192.193 Valve installation in plastic pipe. Each valve installed in plastic pipe must be designed so as to protect the plastic material against excessive torsional or shearing loads when the valve...

  6. 46 CFR 78.17-5 - Valves and closing appliances.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 3 2010-10-01 2010-10-01 false Valves and closing appliances. 78.17-5 Section 78.17-5..., Drills, and Inspections § 78.17-5 Valves and closing appliances. (a) It shall be the duty of the master to see that all valves, including cross connecting valves where fitted, and other appliances such as...

  7. A novel implantable glaucoma valve using ferrofluid.

    Directory of Open Access Journals (Sweden)

    Eleftherios I Paschalis

    Full Text Available PURPOSE: To present a novel design of an implantable glaucoma valve based on ferrofluidic nanoparticles and to compare it with a well-established FDA approved valve. SETTING: Massachusetts Eye & Ear Infirmary, Boston, USA. METHODS: A glaucoma valve was designed using soft lithography techniques utilizing a water-immiscible magnetic fluid (ferrofluid as a pressure-sensitive barrier to aqueous flow. Two rare earth micro magnets were used to calibrate the opening and closing pressure. In-vitro flow measurements were performed to characterize the valve and to compare it to Ahmed™ glaucoma valve. The reliability and predictability of the new valve was verified by pressure/flow measurements over a period of three months and X-ray diffraction (XRD analysis over a period of eight weeks. In vivo assessment was performed in three rabbits. RESULTS: In the in vitro experiments, the opening and closing pressures of the valve were 10 and 7 mmHg, respectively. The measured flow/pressure response was linearly proportional and reproducible over a period of three months (1.8 µl/min at 12 mmHg; 4.3 µl/min at 16 mmHg; 7.6 µl/min at 21 mmHg. X-ray diffraction analysis did not show oxidization of the ferrofluid when exposed to water or air. Preliminary in vivo results suggest that the valve is biocompatible and can control the intraocular pressure in rabbits. CONCLUSIONS: The proposed valve utilizes ferrofluid as passive, tunable constriction element to provide highly predictable opening and closing pressures while maintaining ocular tone. The ferrofluid maintained its magnetic properties in the aqueous environment and provided linear flow to pressure response. Our in-vitro tests showed reliable and reproducible results over a study period of three months. Preliminary in-vivo results were very promising and currently more thorough investigation of this device is underway.

  8. Rotary Valve FY 2016 Highlights

    Energy Technology Data Exchange (ETDEWEB)

    Fitsos, P. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2016-12-07

    The fiscal year started with the Rotary Valve (RV) being reassembled after having crashed in June of 2015. The crash occurred when the RV inner surface contacted the housing. The cause of the crash was never confirmed. No particles were found in the 2.5 thousandths of an inch gap and the filters the helium gas passed through were all clean. There were marks on the bearings that looked like electrostatic discharge as shown below in Figure 1. These marks hadn’t been seen before and there were similar discharge marks on some of the ball bearings. Examples of this were found in a literature search of bearing failures. This leads to a possible cause due to this arcing affecting the rotational accuracy of the bearings driving the RV into the housing.

  9. DISCHARGE VALVE FOR GRANULAR MATERIAL

    Science.gov (United States)

    Stoughton, L.D.; Robinson, S.T.

    1962-05-15

    A gravity-red dispenser or valve is designed for discharging the fueled spherical elements used in a pebble bed reactor. The dispenser consists of an axially movable tube terminating under a hood having side walls with openings. When the tube is moved so that its top edge is above the tops of the side openings the elements will not flow. As the tube is moved downwardly, the elements flow into the hood through the side openings and over the top edge into the tube at an increasing rate as the tube is lowered further. The tube is spaced at all times from the hood and side walls a distance greater than the diameter of the largest element to prevent damaging of the elements when the dispenser is closed to flow. (AEC)

  10. [Late endophthalmitis following Ahmed valve].

    Science.gov (United States)

    del-Hierro-Zarzuelo, A; Vico-Ruiz, E; Martínez-de-la-Casa, J M; García-Feijoó, J; Castillo-Martín, A; García-Sánchez, J

    2005-11-01

    A 71-year-old woman with a history of aphakic glaucoma underwent implantation of an Ahmed valve and scleral grafting in her right eye. Postoperative visual acuity was 0.5 and intraocular pressure was 12 mmHg during treatment with brimonidine tartrate (0.2%). Nine months after implantation she suffered a conjunctival infection which was treated with hygienic measures and topical antibiotic therapy. Four days later, she developed an endophthalmitis which was treated with topical, intravitreous and intravenous vancomycin and ceftazidime. The Ahmed drainage implant was replaced at 72 hours. Laboratory culture yielded Haemophilus influenzae. Four days later, the eye was enucleated. Endophthalmitis is an uncommon complication of glaucoma drainage implant surgery. Exposure of the drainage tube represents the greatest risk factor for this condition. Removal of the implant in the first 24 hours is recommended if a good visual prognosis is to be achieved.

  11. Acoustic emission testing of piston check valves

    International Nuclear Information System (INIS)

    Stewart, D.L.

    1994-01-01

    Based on test experience at Comanche Peak Unit 1, an acoustic emission data evaluation matrix for piston check valves has been developed. The degradations represented in this matrix were selected based on Edwards piston check valve failure data reported in the Nuclear Plant Reliability Data System. Evidence to support this matrix was collected from site test data on a variety of valve types. Although still under refinement, the matrix provides three major attributes for closure verification, which have proven useful in developing test procedures for inservice testing and preventing unnecessary disassembly

  12. Infective endocarditis following transcatheter aortic valve replacement-

    DEFF Research Database (Denmark)

    Loh, Poay Huan; Bundgaard, Henning; S�ndergaard, Lars

    2013-01-01

    Transcatheter aortic valve replacement (TAVR) can improve the symptoms and prognosis of patients with severe aortic stenosis who, due to a high expected operative risk, would not have otherwise been treated surgically. If these patients develop prosthetic valve endocarditis, their presentations may...... be atypical causing a delay in the diagnosis and treatment. The management is also complicated by their comorbidities, and surgical treatment may not be feasible leading to a significant morbidity and mortality. We describe a case of an 85-year-old man with TAVI prosthetic valve endocarditis successfully...

  13. Computed Flow Through An Artificial Heart Valve

    Science.gov (United States)

    Rogers, Stewart E.; Kwak, Dochan; Kiris, Cetin; Chang, I-Dee

    1994-01-01

    Report discusses computations of blood flow through prosthetic tilting disk valve. Computational procedure developed in simulation used to design better artificial hearts and valves by reducing or eliminating following adverse flow characteristics: large pressure losses, which prevent hearts from working efficiently; separated and secondary flows, which causes clotting; and high turbulent shear stresses, which damages red blood cells. Report reiterates and expands upon part of NASA technical memorandum "Computed Flow Through an Artificial Heart and Valve" (ARC-12983). Also based partly on research described in "Numerical Simulation of Flow Through an Artificial Heart" (ARC-12478).

  14. Propionibacterium acnes endophthalmitis in Ahmed glaucoma valve.

    Science.gov (United States)

    Gutiérrez-Díaz, E; Montero-Rodríguez, M; Mencía-Gutiérrez, E; Fernández-González, M C; Pérez-Blázquez, E

    2001-01-01

    To report a case of Propionibacterium acnes endophthalmitis in a patient with an Ahmed glaucoma valve. A nine-year-old boy with bilateral congenital glaucoma, with an Ahmed glaucoma valve implanted in the left eye, had recurrent conjunctival dehiscence and endophthalmitis. Vitreous cultures demonstrated the presence of Propionibacterium acnes. This is the first reported case of Propionibacterium acnes endophthalmitis in an Ahmed glaucoma valve and the second one in a glaucoma drainage device. We strongly recommend using a patch graft to prevent and treat tube exposure. Conjunctival grafts may be useful to close the conjunctiva when there is marked scarring to prevent patch exposure and melting or extrusion.

  15. Can contrast-enhanced multi-detector computed tomography replace transesophageal echocardiography for the detection of thrombogenic milieu and thrombi in the left atrial appendage. A prospective study with 124 patients; Kann die kontrastmittelverstaerkte Mehrzeilen-Computertomografie die transoesophageale Echokardiografie bei der Detektion von thrombogenem Milieu und Vorhofohrthromben ersetzen? Eine prospektive Studie mit 124 Patienten

    Energy Technology Data Exchange (ETDEWEB)

    Homsi, R.; Luetkens, J.A.; Schild, H.H.; Naehle, C.P. [Bonn Univ. (Germany). Dept. of Radiology; Nath, B. [SHG-KLliniken Voelklingen (Germany). Dept. of Medicine I - Cardiology; Schwab, J.O. [Bonn Univ. (Germany). Dept. of Medicine I - Cardiology

    2016-01-15

    To assess the diagnostic value of contrast-enhanced multi-detector computed tomography (MD-CT) for identifying patients with left atrial appendage (LAA) thrombus or circulatory stasis. 124 patients with a history of atrial fibrillation and/or cerebral ischemia (83 men, mean age 58.6 ± 12.4 years) and with a clinical indication for MD-CT of the heart and for transesophageal echocardiography (TEE) were included in the study. LAA thrombus or thrombogenic milieu was visually identified in TEE and MD-CT. In addition, MD-CT was analyzed quantitatively measuring the Hounsfield units (HU) of the left atrium (LA), the LAA and the ascending aorta (AA), and calculating the HU ratios LAA/AA (HU [LAA/AA]) und LAA/LA (HU [LAA/LA]). Sensitivity, specificity, negative predictive value (NPV) and positive predictive value (PPV) were calculated. The prevalence of a thrombus or thrombogenic milieu as assessed by TEE was 21.8 %. The HU ratio was lower in patients with thrombus or thrombogenic milieu (HU [LAA/AA]: 0.590 ± 0.248 vs. 0.909 ± 0.141; p < 0.001 und HU [LAA/LA] 0.689 ± 0.366 vs. 1.082 ± 0.228; p < 0.001). For the diagnosis of thrombus or a thrombogenic milieu, visual analysis yielded a sensitivity of 81.5 %, a specificity of 96.9 %, a PPV of 87.5 % and a NPV of 95.2 %. By combining visual and quantitative analysis with one criterion being positive, the specificity decreased to 91.8 %, the sensitivity to 77.8 %, the PPV to 72.4 %, and the NPV to 94.9 %. Visual analysis of the LAA in the evaluation of thrombus or thrombogenic milieu yields a high NPV of 95.1 % and may especially be useful to rule out LAA thrombi in patients with contraindications for TEE. Additional calculation of HU ratios did not improve the diagnostic performance of MD-CT.

  16. Recurrent pannus formation causing prosthetic aortic valve dysfunction: Is excision without valve re-replacement applicable?

    Directory of Open Access Journals (Sweden)

    Darwazah Ahmad K

    2012-06-01

    Full Text Available Abstract Prosthetic valve dysfunction at aortic position is commonly caused by pannus formation. The exact etiology is not known. It arises from ventricular aspect of the prosthesis encroaching its leaflets causing stenosis or it may remain localized causing left ventricular outflow tract obstruction without affecting valve function. The difference in location entails different approaches in management. Such a pathology requires surgical excision of the pannus with or without valve re-replacement. A recurrent pannus was observed in a female patient who needed repeated surgical intervention to excise a localized pannus without re-replacement of a well functioning prosthetic valve. Management of our case presents several questions, whether recurrence of pannus is caused by sparing the prosthetic valve, is it simply an exaggeration of an inflammatory healing process in certain individuals or is it ideal to re-replace the valve despite a well preserved function.

  17. Recurrent pannus formation causing prosthetic aortic valve dysfunction: is excision without valve re-replacement applicable?

    Science.gov (United States)

    Darwazah, Ahmad K

    2012-06-29

    Prosthetic valve dysfunction at aortic position is commonly caused by pannus formation. The exact etiology is not known. It arises from ventricular aspect of the prosthesis encroaching its leaflets causing stenosis or it may remain localized causing left ventricular outflow tract obstruction without affecting valve function.The difference in location entails different approaches in management. Such a pathology requires surgical excision of the pannus with or without valve re-replacement.A recurrent pannus was observed in a female patient who needed repeated surgical intervention to excise a localized pannus without re-replacement of a well functioning prosthetic valve.Management of our case presents several questions, whether recurrence of pannus is caused by sparing the prosthetic valve, is it simply an exaggeration of an inflammatory healing process in certain individuals or is it ideal to re-replace the valve despite a well preserved function.

  18. Thermally induced pressure locking of gate valves: A survey of valve bonnet pressurization rates

    International Nuclear Information System (INIS)

    Ezekoye, L.I.; Moore, W.E.

    1996-01-01

    Closed, water filled gate valves run the risk of becoming pressurized due to heat input from the environment or from adjacent connected piping. Thermal pressurization of gate valve bonnets may lead to the valves failing to open on demand and can even induce structural failure of valves. This paper presents an analytical prediction of the pressurization rate of a closed pressure vessel subject to uniform heating which may be considered as an upper bound to the pressurization rate that may occur in the field. Then actual valve experiences described in the literature are reviewed to determine the expected pressurization rate in existing hardware designs. A statistical approach is applied to reconcile the differing pressurization rates reported in the literature and determine a rate that can be applied in valve evaluations. The limitations of the reconciled rate are discussed

  19. Fully coupled fluid-structure interaction model of reed valves in a multi-cylinder reciprocating piston compressor

    Science.gov (United States)

    Xie, F.; Nieter, J.; Lifson, A.; Reba, R.; Sishtla, V.

    2017-08-01

    For years compressor researchers have tried to account for the fluid interaction effect of the working fluid on valve motion in displacement compressors. In recent years, the computing capacities and available CFD and FEA programs have allowed fully coupled interaction of fluids and moving structures to be modelled more comprehensively. This paper describes our experience and results from developing a model of a multi-cylinder reciprocating piston compressor with suction and discharge valve systems that are fully coupled with the pressure pulsation in the adjacent plenum. Valve dynamics are captured by the model that affects compressor performance. The results show that higher running speed causes more discharge valve delay on closing due to higher pressure pulsation in discharge plenum. The acoustic property of the discharge plenum as it relates to valve motion is studied by the developed cost-effective standalone model.

  20. Dysfunction of an On-X Heart Valve by Pannus.

    Science.gov (United States)

    Abad, Cipriano; Urso, Stefano; Gomez, Elsa; De la Vega, Maria

    2016-09-01

    A 68-year-old woman with a history of previous double-valve replacement with On-X mechanical heart valves presented with clinical, echocardiographic and cardiac catheterization signs of obstruction of the On-X tricuspid heart valve prosthesis. The patient was successfully reoperated, but at surgery the valve was seen to be invaded by an abnormal overgrowth of pannus that blocked one of the leaflets. A small amount of non-obstructive fresh thrombus was also observed. The valve was successfully replaced with a biological heart valve prosthesis. The patient was discharged home, and is doing well four months after the operation, when echocardiography demonstrated normal function in the tricuspid valve. The present case represents the first ever report of pannus formation and subsequent dysfunction in an On-X heart valve, and also the first case of tricuspid valve malfunction and obstruction using this type of heart valve substitute.