WorldWideScience

Sample records for unit preflight calibration

  1. TIMED solar EUV experiment: preflight calibration results for the XUV photometer system

    Science.gov (United States)

    Woods, Thomas N.; Rodgers, Erica M.; Bailey, Scott M.; Eparvier, Francis G.; Ucker, Gregory J.

    1999-10-01

    The Solar EUV Experiment (SEE) on the NASA Thermosphere, Ionosphere, and Mesosphere Energetics and Dynamics (TIMED) mission will measure the solar vacuum ultraviolet (VUV) spectral irradiance from 0.1 to 200 nm. To cover this wide spectral range two different types of instruments are used: a grating spectrograph for spectra between 25 and 200 nm with a spectral resolution of 0.4 nm and a set of silicon soft x-ray (XUV) photodiodes with thin film filters as broadband photometers between 0.1 and 35 nm with individual bandpasses of about 5 nm. The grating spectrograph is called the EUV Grating Spectrograph (EGS), and it consists of a normal- incidence, concave diffraction grating used in a Rowland spectrograph configuration with a 64 X 1024 array CODACON detector. The primary calibrations for the EGS are done using the National Institute for Standards and Technology (NIST) Synchrotron Ultraviolet Radiation Facility (SURF-III) in Gaithersburg, Maryland. In addition, detector sensitivity and image quality, the grating scattered light, the grating higher order contributions, and the sun sensor field of view are characterized in the LASP calibration laboratory. The XUV photodiodes are called the XUV Photometer System (XPS), and the XPS includes 12 photodiodes with thin film filters deposited directly on the silicon photodiodes' top surface. The sensitivities of the XUV photodiodes are calibrated at both the NIST SURF-III and the Physikalisch-Technische Bundesanstalt (PTB) electron storage ring called BESSY. The other XPS calibrations, namely the electronics linearity and field of view maps, are performed in the LASP calibration laboratory. The XPS and solar sensor pre-flight calibration results are primarily discussed as the EGS calibrations at SURF-III have not yet been performed.

  2. The pre-flight calibration setup of the instrument SIMBIO-SYS onboard the mission BepiColombo

    Science.gov (United States)

    Poulet, F.; Rodriguez-Ferreira, J.; Arondel, A.; Dassas, K.; Eng, P.; Lami, P.; Langevin, Y.; Longval, Y.; Pradel, P.; Dami, M.

    2015-11-01

    BepiColombo, an European Space Agency (ESA) mission being conducted in cooperation with the Japan space agency, will explore Mercury with a set of eleven instruments onboard the spacecraft Mercury Planetary Orbiter (MPO). Among them, SIMBIO-SYS (Spectrometers and Imagers for MPO BepiColombo Integrated Observatory SYStem) is a complex instrument that will provide images and spectra in the 400-2000 nm wavelength range of the entire surface of Mercury. Pre-flight calibration of the SYMBIO-SYS instrument is mandatory for reliable scientific interpretation of images and spectra returned from the planet Mercury. This paper presents the calibration device designed and implemented for the specific requirements of this instrument. It mainly consists of a thermal vacuum chamber simulating the space environment, an optical bench collecting calibration sources and optical elements that simulate the conditions of Mercury observations, mechanical interfaces used for positioning the three channels inside the vacuum chamber, thermal interfaces to explore the operating temperatures, computer interfaces that allow to communicate with both the instrument and the calibration elements and synchronize the calibrations sequences with the status of the calibration device. As the major goal is the characterization of the radiometric performances of the three channels of SIMBIO-SYS, radiometric performances of the test setup evaluated by simulations and measurements are emphasized.

  3. CMOS detectors: lessons learned during the STC stereo channel preflight calibration

    Science.gov (United States)

    Simioni, E.; De Sio, A.; Da Deppo, V.; Naletto, G.; Cremonese, G.

    2017-09-01

    The Stereo Camera (STC), mounted on-board the BepiColombo spacecraft, will acquire in push frame stereo mode the entire surface of Mercury. STC will provide the images for the global three-dimensional reconstruction of the surface of the innermost planet of the Solar System. The launch of BepiColombo is foreseen in 2018. STC has an innovative optical system configuration, which allows good optical performances with a mass and volume reduction of a factor two with respect to classical stereo camera approach. In such a telescope, two different optical paths inclined of +/-20°, with respect to the nadir direction, are merged together in a unique off axis path and focused on a single detector. The focal plane is equipped with a 2k x 2k hybrid Si-PIN detector, based on CMOS technology, combining low read-out noise, high radiation hardness, compactness, lack of parasitic light, capability of snapshot image acquisition and short exposure times (less than 1 ms) and small pixel size (10 μm). During the preflight calibration campaign of STC, some detector spurious effects have been noticed. Analyzing the images taken during the calibration phase, two different signals affecting the background level have been measured. These signals can reduce the detector dynamics down to a factor of 1/4th and they are not due to dark current, stray light or similar effects. In this work we will describe all the features of these unwilled effects, and the calibration procedures we developed to analyze them.

  4. The Preflight Photometric Calibration of the Extreme-Ultraviolet Imaging Telescope EIT

    Science.gov (United States)

    Dere, K. P.; Moses, J. D.; Delaboudiniere, J. -P.; Brunaud, J.; Carabetian, C.; Hochedez, J. -F.; Song, X. Y.; Catura, R. C.; Clette, F.; Defise, J. -M.

    2000-01-01

    This paper presents the preflight photometric calibration of the Extreme-ultraviolet Imaging Telescope (EIT) aboard the Solar and Heliospheric Observatory (SOHO). The EIT consists of a Ritchey-Chretien telescope with multilayer coatings applied to four quadrants of the primary and secondary mirrors, several filters and a backside-thinned CCD detector. The quadrants of the EIT optics were used to observe the Sun in 4 wavelength bands that peak near 171, 195, 284, and 304 . Before the launch of SOHO, the EIT mirror reflectivities, the filter transmissivities and the CCD quantum efficiency were measured and these values are described here. The instrumental throughput in terms of an effective area is presented for each of the various mirror quadrant and filter wheel combinations. The response to a coronal plasma as a function of temperature is also determined and the expected count rates are compared to the count rates observed in a coronal hole, the quiet Sun and an active region.

  5. Ørsted Pre-Flight Magnetometer Calibration Mission

    DEFF Research Database (Denmark)

    Risbo, T.; Brauer, Peter; Merayo, José M.G.

    2003-01-01

    and the overall calibration results are given. The temperature calibrations are explained and reported on. The overall calibration model standard deviation is about 100 pT rms. Comparisons with the later in-flight calibrations show that, except for the unknown satellite offsets, an agreement within 4 n...

  6. The Mars Science Laboratory Curiosity rover Mastcam instruments: Preflight and in-flight calibration, validation, and data archiving

    Science.gov (United States)

    Bell, J. F.; Godber, A.; McNair, S.; Caplinger, M. A.; Maki, J. N.; Lemmon, M. T.; Van Beek, J.; Malin, M. C.; Wellington, D.; Kinch, K. M.; Madsen, M. B.; Hardgrove, C.; Ravine, M. A.; Jensen, E.; Harker, D.; Anderson, R. B.; Herkenhoff, K. E.; Morris, R. V.; Cisneros, E.; Deen, R. G.

    2017-07-01

    The NASA Curiosity rover Mast Camera (Mastcam) system is a pair of fixed-focal length, multispectral, color CCD imagers mounted 2 m above the surface on the rover's remote sensing mast, along with associated electronics and an onboard calibration target. The left Mastcam (M-34) has a 34 mm focal length, an instantaneous field of view (IFOV) of 0.22 mrad, and a FOV of 20° × 15° over the full 1648 × 1200 pixel span of its Kodak KAI-2020 CCD. The right Mastcam (M-100) has a 100 mm focal length, an IFOV of 0.074 mrad, and a FOV of 6.8° × 5.1° using the same detector. The cameras are separated by 24.2 cm on the mast, allowing stereo images to be obtained at the resolution of the M-34 camera. Each camera has an eight-position filter wheel, enabling it to take Bayer pattern red, green, and blue (RGB) "true color" images, multispectral images in nine additional bands spanning 400-1100 nm, and images of the Sun in two colors through neutral density-coated filters. An associated Digital Electronics Assembly provides command and data interfaces to the rover, 8 Gb of image storage per camera, 11 bit to 8 bit companding, JPEG compression, and acquisition of high-definition video. Here we describe the preflight and in-flight calibration of Mastcam images, the ways that they are being archived in the NASA Planetary Data System, and the ways that calibration refinements are being developed as the investigation progresses on Mars. We also provide some examples of data sets and analyses that help to validate the accuracy and precision of the calibration.

  7. The Mars Science Laboratory Curiosity rover Mastcam instruments: Preflight and in-flight calibration, validation, and data archiving

    Science.gov (United States)

    Bell, James F.; Godber, A.; McNair, S.; Caplinger, M.A.; Maki, J.N.; Lemmon, M.T.; Van Beek, J.; Malin, M.C.; Wellington, D.; Kinch, K.M.; Madsen, M.B.; Hardgrove, C.; Ravine, M.A.; Jensen, E.; Harker, D.; Anderson, Ryan; Herkenhoff, Kenneth E.; Morris, R.V.; Cisneros, E.; Deen, R.G.

    2017-01-01

    The NASA Curiosity rover Mast Camera (Mastcam) system is a pair of fixed-focal length, multispectral, color CCD imagers mounted ~2 m above the surface on the rover's remote sensing mast, along with associated electronics and an onboard calibration target. The left Mastcam (M-34) has a 34 mm focal length, an instantaneous field of view (IFOV) of 0.22 mrad, and a FOV of 20° × 15° over the full 1648 × 1200 pixel span of its Kodak KAI-2020 CCD. The right Mastcam (M-100) has a 100 mm focal length, an IFOV of 0.074 mrad, and a FOV of 6.8° × 5.1° using the same detector. The cameras are separated by 24.2 cm on the mast, allowing stereo images to be obtained at the resolution of the M-34 camera. Each camera has an eight-position filter wheel, enabling it to take Bayer pattern red, green, and blue (RGB) “true color” images, multispectral images in nine additional bands spanning ~400–1100 nm, and images of the Sun in two colors through neutral density-coated filters. An associated Digital Electronics Assembly provides command and data interfaces to the rover, 8 Gb of image storage per camera, 11 bit to 8 bit companding, JPEG compression, and acquisition of high-definition video. Here we describe the preflight and in-flight calibration of Mastcam images, the ways that they are being archived in the NASA Planetary Data System, and the ways that calibration refinements are being developed as the investigation progresses on Mars. We also provide some examples of data sets and analyses that help to validate the accuracy and precision of the calibration

  8. A verified technique for calibrating space solar cells

    Science.gov (United States)

    Anspaugh, Bruce

    1987-01-01

    Solar cells have been flown on high-altitude balloons for over 24 years, to produce solar cell standards that can be used to set the intensity of solar simulators. The events of a typical balloon calibration flight are reported. These are: the preflight events, including the preflight cell measurements and the assembly of the flight cells onto the solar tracker; the activities at the National Scientific Balloon Facility in Palestine, Texas, including the preflight calibrations, the mating of the tracker and cells onto the balloon, preparations for launch, and the launch; the payload recovery, which includes tracking the balloon by aircraft, terminating the flight, and retrieving the payload. In 1985, the cells flow on the balloon were also flown on a shuttle flight and measured independently. The two measurement methods are compared and shown to agree within 1 percent.

  9. Calibration-measurement unit for the automation of vector network analyzer measurements

    Directory of Open Access Journals (Sweden)

    I. Rolfes

    2008-05-01

    Full Text Available With the availability of multi-port vector network analyzers, the need for automated, calibrated measurement facilities increases. In this contribution, a calibration-measurement unit is presented which realizes a repeatable automated calibration of the measurement setup as well as a user-friendly measurement of the device under test (DUT. In difference to commercially available calibration units, which are connected to the ports of the vector network analyzer preceding a measurement and which are then removed so that the DUT can be connected, the presented calibration-measurement unit is permanently connected to the ports of the VNA for the calibration as well as for the measurement of the DUT. This helps to simplify the calibrated measurement of complex scattering parameters. Moreover, a full integration of the calibration unit into the analyzer setup becomes possible. The calibration-measurement unit is based on a multiport switch setup of e.g. electromechanical relays. Under the assumption of symmetry of a switch, on the one hand the unit realizes the connection of calibration standards like one-port reflection standards and two-port through connections between different ports and on the other hand it enables the connection of the DUT. The calibration-measurement unit is applicable for two-port VNAs as well as for multiport VNAs. For the calibration of the unit, methods with completely known calibration standards like SOLT (short, open, load, through as well as self-calibration procedures like TMR or TLR can be applied.

  10. Trajectory Reconstruction and Uncertainty Analysis Using Mars Science Laboratory Pre-Flight Scale Model Aeroballistic Testing

    Science.gov (United States)

    Lugo, Rafael A.; Tolson, Robert H.; Schoenenberger, Mark

    2013-01-01

    As part of the Mars Science Laboratory (MSL) trajectory reconstruction effort at NASA Langley Research Center, free-flight aeroballistic experiments of instrumented MSL scale models was conducted at Aberdeen Proving Ground in Maryland. The models carried an inertial measurement unit (IMU) and a flush air data system (FADS) similar to the MSL Entry Atmospheric Data System (MEADS) that provided data types similar to those from the MSL entry. Multiple sources of redundant data were available, including tracking radar and on-board magnetometers. These experimental data enabled the testing and validation of the various tools and methodologies that will be used for MSL trajectory reconstruction. The aerodynamic parameters Mach number, angle of attack, and sideslip angle were estimated using minimum variance with a priori to combine the pressure data and pre-flight computational fluid dynamics (CFD) data. Both linear and non-linear pressure model terms were also estimated for each pressure transducer as a measure of the errors introduced by CFD and transducer calibration. Parameter uncertainties were estimated using a "consider parameters" approach.

  11. Otolith tilt-translation reinterpretation following prolonged weightlessness - Implications for preflight training

    Science.gov (United States)

    Parker, D. E.; Reschke, M. F.; Arrott, A. P.; Homick, J. L.; Lichtenberg, B. K.

    1985-01-01

    Observations with three astronauts yielded two major findings. First, perceived self-motion during sinusoidal roll differed immediately postflight from preflight. Between 70 and 150 min after landing, roll was perceived primarily as linear translation. Secondly, more horizontal eye movement was elicited by roll simulation immediately postflight relative to both preflight and later postflight observations. These results support an 'otolith tilt-translation reinterpretation' hypothesis, which has clear implications for understanding astronaut reports of space motion sickness during the early period of orbital flight. A proposal for 'prophylactic adaptation training' which may provide preflight adaptation to weightlessness, derives from this reearch.

  12. Calibration is action specific but perturbation of perceptual units is not.

    Science.gov (United States)

    Pan, Jing S; Coats, Rachel O; Bingham, Geoffrey P

    2014-02-01

    G. P. Bingham and C. C. Pagano (1998, The necessity of a perception/action approach to definite distance perception: Monocular distance perception to guide reaching. Journal of Experimental Psychology: Human Perception and Performance, 24, 145-168) argued that metric space perception should be investigated using relevant action measures because calibration is an intrinsic component of perception/action that yields accurate targeted actions. They described calibration as a mapping from embodied units of perception to embodied units of action. This mapping theory yields a number of predictions. We tested two of them. The first prediction is that calibration should be action specific because what is calibrated is a mapping from perceptual units to a unit of action. Thus, calibration does not generalize to other actions. This prediction is consistent with the "action-specific approach" to calibration (D. R. Proffitt, 2008, An action specific approach to spatial perception. In R. L. Klatzky, B. MacWhinney, & M. Behrmann (Eds.), Embodiment, ego-space and action (pp. 179-202). New York, NY: Psychology Press.). The second prediction is that a change in perceptual units should generalize to all relevant actions that are guided using that perceptual information. The same perceptual units can be mapped to different actions. Change in the unit affects all relevant actions. This prediction is consistent with the "general purpose perception approach" (J. M. Loomis & J. W. Philbeck, 2008, Measuring spatial perception with spatial updating and action. In R. L. Klatzky, B. MacWhinney, & M. Behrmann (Eds.), Embodiment, ego-space and action (pp. 1-43). New York, NY: Psychology Press). In Experiment 1, two targeted actions, throwing and extended reaching were tested to determine if they were comparable in precision and in response to distorted calibration. They were. Comparing these actions, the first prediction was tested in Experiment 2 and confirmed. The second prediction was

  13. ESTABLISHING BRDF CALIBRATION CAPABILITIES THROUGH SHORTWAVE INFRARED

    OpenAIRE

    Georgiev, Georgi T.; Butler, James J.; Thome, Kurt; Cooksey, Catherine; Ding, Leibo

    2017-01-01

    Satellite instruments operating in the reflective solar wavelength region require accurate and precise determination of the Bidirectional Reflectance Distribution Functions (BRDFs) of the laboratory and flight diffusers used in their pre-flight and on-orbit calibrations. This paper advances that initial work and presents a comparison of spectral Bidirectional Reflectance Distribution Function (BRDF) and Directional Hemispherical Reflectance (DHR) of Spectralon*, a common material for laborato...

  14. An Optimal Calibration Method for a MEMS Inertial Measurement Unit

    Directory of Open Access Journals (Sweden)

    Bin Fang

    2014-02-01

    Full Text Available An optimal calibration method for a micro-electro-mechanical inertial measurement unit (MIMU is presented in this paper. The accuracy of the MIMU is highly dependent on calibration to remove the deterministic errors of systematic errors, which also contain random errors. The overlapping Allan variance is applied to characterize the types of random error terms in the measurements. The calibration model includes package misalignment error, sensor-to-sensor misalignment error and bias, and a scale factor is built. The new concept of a calibration method, which includes a calibration scheme and a calibration algorithm, is proposed. The calibration scheme is designed by D-optimal and the calibration algorithm is deduced by a Kalman filter. In addition, the thermal calibration is investigated, as the bias and scale factor varied with temperature. The simulations and real tests verify the effectiveness of the proposed calibration method and show that it is better than the traditional method.

  15. A new Cassegrain calibration lamp unit for the Blanco Telescope

    Science.gov (United States)

    Points, S. D.; James, D. J.; Tighe, R.; Montané, A.; David, N.; Martínez, M.

    2016-08-01

    The f/8 RC-Cassegrain Focus of the Blanco Telescope at Cerro Tololo Inter-American Observatory, hosts two new instruments: COSMOS, a multi-object spectrograph in the visible wavelength range (350 - 1030nm), and ARCoIRIS, a NIR cross-dispersed spectrograph featuring 6 spectral orders spanning 0.8 - 2.45μm. Here we describe a calibration lamp unit designed to deliver the required illumination at the telescope focal plane for both instruments. These requirements are: (1) an f/8 beam of light covering a spot of 92mm diameter (or 10 arcmin) for a wavelength range of 0.35μm through 2.5μm and (2) no saturation of flat-field calibrations for the minimal exposure times permitted by each instrument, and (3) few saturated spectral lines when using the wavelength calibration lamps for the instruments. To meet these requirements this unit contains an adjustable quartz halogen lamp for flat-field calibrations, and one hollow cathode lamp and four penray lamps for wavelength calibrations. The wavelength calibration lamps are selected to provide optimal spectral coverage for the instrument mounted and can be used individually or in sets. The device designed is based on an 8-inch diameter integrating sphere, the output of which is optimized to match the f/8 calibration input delivery system which is a refractive system based on fused-silica lenses. We describe the optical design, the opto-mechanical design, the electronic control and give results of the performance of the system.

  16. Systematic Calibration for Ultra-High Accuracy Inertial Measurement Units

    Directory of Open Access Journals (Sweden)

    Qingzhong Cai

    2016-06-01

    Full Text Available An inertial navigation system (INS has been widely used in challenging GPS environments. With the rapid development of modern physics, an atomic gyroscope will come into use in the near future with a predicted accuracy of 5 × 10−6°/h or better. However, existing calibration methods and devices can not satisfy the accuracy requirements of future ultra-high accuracy inertial sensors. In this paper, an improved calibration model is established by introducing gyro g-sensitivity errors, accelerometer cross-coupling errors and lever arm errors. A systematic calibration method is proposed based on a 51-state Kalman filter and smoother. Simulation results show that the proposed calibration method can realize the estimation of all the parameters using a common dual-axis turntable. Laboratory and sailing tests prove that the position accuracy in a five-day inertial navigation can be improved about 8% by the proposed calibration method. The accuracy can be improved at least 20% when the position accuracy of the atomic gyro INS can reach a level of 0.1 nautical miles/5 d. Compared with the existing calibration methods, the proposed method, with more error sources and high order small error parameters calibrated for ultra-high accuracy inertial measurement units (IMUs using common turntables, has a great application potential in future atomic gyro INSs.

  17. 14 CFR 417.121 - Safety critical preflight operations.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 4 2010-01-01 2010-01-01 false Safety critical preflight operations. 417.121 Section 417.121 Aeronautics and Space COMMERCIAL SPACE TRANSPORTATION, FEDERAL AVIATION... surveillance. A launch operator must implement its hazard area surveillance and clearance plan, of § 417.111(j...

  18. An Improved Calibration Method for Hydrazine Monitors for the United States Air Force

    Energy Technology Data Exchange (ETDEWEB)

    Korsah, K

    2003-07-07

    This report documents the results of Phase 1 of the ''Air Force Hydrazine Detector Characterization and Calibration Project''. A method for calibrating model MDA 7100 hydrazine detectors in the United States Air Force (AF) inventory has been developed. The calibration system consists of a Kintek 491 reference gas generation system, a humidifier/mixer system which combines the dry reference hydrazine gas with humidified diluent or carrier gas to generate the required humidified reference for calibrations, and a gas sampling interface. The Kintek reference gas generation system itself is periodically calibrated using an ORNL-constructed coulometric titration system to verify the hydrazine concentration of the sample atmosphere in the interface module. The Kintek reference gas is then used to calibrate the hydrazine monitors. Thus, coulometric titration is only used to periodically assess the performance of the Kintek reference gas generation system, and is not required for hydrazine monitor calibrations. One advantage of using coulometric titration for verifying the concentration of the reference gas is that it is a primary standard (if used for simple solutions), thereby guaranteeing, in principle, that measurements will be traceable to SI units (i.e., to the mole). The effect of humidity of the reference gas was characterized by using the results of concentrations determined by coulometric titration to develop a humidity correction graph for the Kintek 491 reference gas generation system. Using this calibration method, calibration uncertainty has been reduced by 50% compared to the current method used to calibrate hydrazine monitors in the Air Force inventory and calibration time has also been reduced by more than 20%. Significant findings from studies documented in this report are the following: (1) The Kintek 491 reference gas generation system (generator, humidifier and interface module) can be used to calibrate hydrazine detectors. (2) The

  19. BRDF Calibration of Sintered PTFE in the SWIR

    Science.gov (United States)

    Georgiev, Georgi T.; Butler, James J.

    2009-01-01

    Satellite instruments operating in the reflective solar wavelength region often require accurate and precise determination of the Bidirectional Reflectance Distribution Function (BRDF) of laboratory-based diffusers used in their pre-flight calibrations and ground-based support of on-orbit remote sensing instruments. The Diffuser Calibration Facility at NASA's Goddard Space Flight Center is a secondary diffuser calibration standard after NEST for over two decades, providing numerous NASA projects with BRDF data in the UV, Visible and the NIR spectral regions. Currently the Diffuser Calibration Facility extended the covered spectral range from 900 nm up to 1.7 microns. The measurements were made using the existing scatterometer by replacing the Si photodiode based receiver with an InGaAs-based one. The BRDF data was recorded at normal incidence and scatter zenith angles from 10 to 60 deg. Tunable coherent light source was setup. Broadband light source application is under development. Gray-scale sintered PTFE samples were used at these first trials, illuminated with P and S polarized incident light. The results are discussed and compared to empirically generated BRDF data from simple model based on 8 deg directional/hemispherical measurements.

  20. Inflight and Preflight Detection of Pitot Tube Anomalies

    Science.gov (United States)

    Mitchell, Darrell W.

    2014-01-01

    The health and integrity of aircraft sensors play a critical role in aviation safety. Inaccurate or false readings from these sensors can lead to improper decision making, resulting in serious and sometimes fatal consequences. This project demonstrated the feasibility of using advanced data analysis techniques to identify anomalies in Pitot tubes resulting from blockage such as icing, moisture, or foreign objects. The core technology used in this project is referred to as noise analysis because it relates sensors' response time to the dynamic component (noise) found in the signal of these same sensors. This analysis technique has used existing electrical signals of Pitot tube sensors that result from measured processes during inflight conditions and/or induced signals in preflight conditions to detect anomalies in the sensor readings. Analysis and Measurement Services Corporation (AMS Corp.) has routinely used this technology to determine the health of pressure transmitters in nuclear power plants. The application of this technology for the detection of aircraft anomalies is innovative. Instead of determining the health of process monitoring at a steady-state condition, this technology will be used to quickly inform the pilot when an air-speed indication becomes faulty under any flight condition as well as during preflight preparation.

  1. Evaluation of the need for radiotherapy calibrations in the United States

    International Nuclear Information System (INIS)

    Lanzl, L.H.; Rozenfeld, M.; r.

    1977-01-01

    The custody, maintenance, and development of the United States national standards of measurement, together with the provision of calibration services related to these standards, has been assigned to the National Bureau of Standards (NBS) of the U. S. Department of Commerce. It may be that the present system of promulgation of radiation standards is no longer adequate for the United States. Ionizing radiation is being used increasingly for the treatment of cancer patients. At present, approximately 325,000 new cancer patients per year are given radiation treatment at over 1500 centers. With this growing use of ionizing radiations, an increased awareness of the importance of accurate and precise dosimetry has developed. To determine the adequacy of these dosimetric needs, the AAPM has initiated a national study which is being conducted by means of a mailed questionnaire, and by a limited number of site visits to individuals who are in charge of radiation therapy centers as well as to those who do the actual calibrations

  2. Orbit transfer rocket engine technology program: Automated preflight methods concept definition

    Science.gov (United States)

    Erickson, C. M.; Hertzberg, D. W.

    1991-01-01

    The possibility of automating preflight engine checkouts on orbit transfer engines is discussed. The minimum requirements in terms of information and processing necessary to assess the engine'e integrity and readiness to perform its mission were first defined. A variety of ways for remotely obtaining that information were generated. The sophistication of these approaches varied from a simple preliminary power up, where the engine is fired up for the first time, to the most advanced approach where the sensor and operational history data system alone indicates engine integrity. The critical issues and benefits of these methods were identified, outlined, and prioritized. The technology readiness of each of these automated preflight methods were then rated on a NASA Office of Exploration scale used for comparing technology options for future mission choices. Finally, estimates were made of the remaining cost to advance the technology for each method to a level where the system validation models have been demonstrated in a simulated environment.

  3. A hardware-software system for the automation of verification and calibration of oil metering units secondary equipment

    Science.gov (United States)

    Boyarnikov, A. V.; Boyarnikova, L. V.; Kozhushko, A. A.; Sekachev, A. F.

    2017-08-01

    In the article the process of verification (calibration) of oil metering units secondary equipment is considered. The purpose of the work is to increase the reliability and reduce the complexity of this process by developing a software and hardware system that provides automated verification and calibration. The hardware part of this complex carries out the commutation of the measuring channels of the verified controller and the reference channels of the calibrator in accordance with the introduced algorithm. The developed software allows controlling the commutation of channels, setting values on the calibrator, reading the measured data from the controller, calculating errors and compiling protocols. This system can be used for checking the controllers of the secondary equipment of the oil metering units in the automatic verification mode (with the open communication protocol) or in the semi-automatic verification mode (without it). The peculiar feature of the approach used is the development of a universal signal switch operating under software control, which can be configured for various verification methods (calibration), which allows to cover the entire range of controllers of metering units secondary equipment. The use of automatic verification with the help of a hardware and software system allows to shorten the verification time by 5-10 times and to increase the reliability of measurements, excluding the influence of the human factor.

  4. The Calibration Units of the KM3NeT neutrino telescope

    Science.gov (United States)

    Baret, B.; Keller, P.; Clark, M. Lindsey

    2016-04-01

    KM3NeT is a network of deep-sea neutrino telescopes to be deployed in the Mediterranean Sea that will perform neutrino astronomy and oscillation studies. It consists of three-dimensional arrays of thousands of optical modules that detect the Cherenkov light induced by charged particles resulting from the interaction of a neutrino with the surrounding medium. The performance of the neutrino telescope relies on the precise timing and positioning calibration of the detector elements. Other environmental conditions which may affect light and sound transmission, such as water temperature and salinity, must also be continuously monitored. This contribution describes the technical design of the first Calibration Unit, to be deployed on the French site as part of KM3NeT Phase 1.

  5. The assembly, calibration, and preliminary results from the Colorado high-resolution Echelle stellar spectrograph (CHESS)

    Science.gov (United States)

    Hoadley, Keri; France, Kevin; Nell, Nicholas; Kane, Robert; Schultz, Ted; Beasley, Matthew; Green, James; Kulow, Jen; Kersgaard, Eliot; Fleming, Brian

    2014-07-01

    The Colorado High-resolution Echelle Stellar Spectrograph (CHESS) is a far ultraviolet (FUV) rocket-borne experiment designed to study the atomic-to-molecular transitions within translucent interstellar clouds. CHESS is an objective echelle spectrograph operating at f/12.4 and resolving power of 120,000 over a band pass of 100 - 160 nm. The echelle flight grating is the product of a research and development project with LightSmyth Inc. and was coated at Goddard Space Flight Center (GSFC) with Al+LiF. It has an empirically-determined groove density of 71.67 grooves/mm. At the Center for Astrophysics and Space Astronomy (CASA) at the University of Colorado (CU), we measured the efficiencies of the peak and adjacent dispersion orders throughout the 90 - 165 nm band pass to characterize the behavior of the grating for pre-flight calibrations and to assess the scattered-light behavior. The crossdispersing grating, developed and ruled by Horiba Jobin-Yvon, is a holographically-ruled, low line density (351 grooves/mm), powered optic with a toroidal surface curvature. The CHESS cross-disperser was also coated at GSFC; Cr+Al+LiF was deposited to enhance far-UV efficiency. Results from final efficiency and reflectivity measurements of both optics are presented. We utilize a cross-strip anode microchannel plate (MCP) detector built by Sensor Sciences to achieve high resolution (25 μm spatial resolution) and data collection rates (~ 106 photons/second) over a large format (40mm round, digitized to 8k x 8k) for the first time in an astronomical sounding rocket flight. The CHESS instrument was successfully launched from White Sands Missile Range on 24 May 2014. We present pre-flight sensitivity, effective area calculations, lab spectra and calibration results, and touch on first results and post-flight calibration plans.

  6. STS-30 Pilot Ronald J. Grabe during preflight press conference at JSC

    Science.gov (United States)

    1989-01-01

    During preflight press conference, STS-30 Pilot Ronald J. Grabe answers a question from the news media. The event was held in the JSC Auditorium and Public Affairs Facility Bldg 2 briefing room. STS-30 mission will fly onboard Atlantis, Orbiter Vehicle (OV) 104, and is scheduled for an April 28 liftoff.

  7. The assessment of toxic emission from the engines of the W-3 helicopter in the preflight engine run

    Directory of Open Access Journals (Sweden)

    Cur Krzysztof

    2015-12-01

    Full Text Available Helicopters are used in a variety of tasks like sea, mountain and land search and rescue, civil transportation or military operations. Their amount is growing, their ceiling is usually low, however, the exhaust emission from their engines is not restricted by any directive. That is why an attempt was made to determine the real exhaust emission. The research was done on the W-3 helicopter during the preflight engine run. Based on the results set the standard was created which reflects the working conditions of the engines during the test cycle. The research allowed to establish the unit emission of toxic gasses and then the influence on environment of the engines.

  8. A program on quality assurance and dose calibration for radiation therapy units in Venezuela

    International Nuclear Information System (INIS)

    Padilla, M.C. de; Carrizales, L.; Diaz, J.; Gutt, F.; Cozman, A.

    1996-01-01

    The results of a five year program (1988-90-91-92-93) on quality assurance and dose calibration for 12 Cobalt-60 units from public hospitals, which represents 30% of total radiation therapy units in Venezuela, are presented. The remarkable improvement in the general performance of these units can be seen in the IAEA/WHO Postal TLD Intercomparison results which gave 100% within ± 5% in 1990 and 1992, while 63% in 1990 and 44% in 1992, with errors up to 37% were obtained for the participants not included in the program. The difference between the two groups lead the government to decrete through the Gaceta Oficial de la Republica de Venezuela, Resolution G-1397 on March 3, 1993, the quality assurance and dose calibration programs shall be established for all radiation therapy installations in Venezuela. The project for the standards was developed by the SSDL physicists and it was already approbated by the Health Ministry. It is expected that the Norms will enter into effect by the end of 1994. (author). 14 refs, 1 fig., 3 tabs

  9. A program on quality assurance and dose calibration for radiation therapy units in Venezuela

    Energy Technology Data Exchange (ETDEWEB)

    Padilla, M.C. de; Carrizales, L; Diaz, J; Gutt, F; Cozman, A [Instituto Venezolano de Investigaciones Cientificas, Caracas (Venezuela). Lab. de Calibracion Dosimetrica

    1996-08-01

    The results of a five year program (1988-90-91-92-93) on quality assurance and dose calibration for 12 Cobalt-60 units from public hospitals, which represents 30% of total radiation therapy units in Venezuela, are presented. The remarkable improvement in the general performance of these units can be seen in the IAEA/WHO Postal TLD Intercomparison results which gave 100% within {+-} 5% in 1990 and 1992, while 63% in 1990 and 44% in 1992, with errors up to 37% were obtained for the participants not included in the program. The difference between the two groups lead the government to decrete through the Gaceta Oficial de la Republica de Venezuela, Resolution G-1397 on March 3, 1993, the quality assurance and dose calibration programs shall be established for all radiation therapy installations in Venezuela. The project for the standards was developed by the SSDL physicists and it was already approbated by the Health Ministry. It is expected that the Norms will enter into effect by the end of 1994. (author). 14 refs, 1 fig., 3 tabs.

  10. Poster Abstract: Automatic Calibration of Device Attitude in Inertial Measurement Unit Based Traffic Probe Vehicles

    KAUST Repository

    Mousa, Mustafa; Sharma, Kapil; Claudel, Christian

    2016-01-01

    to replace them with inertial measurement units onboard vehicles, to estimate vehicle location and attitude using inertial data only. While promising, this technology requires one to carefully calibrate the orientation of the device inside the vehicle

  11. Calibration factor or calibration coefficient?

    International Nuclear Information System (INIS)

    Meghzifene, A.; Shortt, K.R.

    2002-01-01

    Full text: The IAEA/WHO network of SSDLs was set up in order to establish links between SSDL members and the international measurement system. At the end of 2001, there were 73 network members in 63 Member States. The SSDL network members provide calibration services to end-users at the national or regional level. The results of the calibrations are summarized in a document called calibration report or calibration certificate. The IAEA has been using the term calibration certificate and will continue using the same terminology. The most important information in a calibration certificate is a list of calibration factors and their related uncertainties that apply to the calibrated instrument for the well-defined irradiation and ambient conditions. The IAEA has recently decided to change the term calibration factor to calibration coefficient, to be fully in line with ISO [ISO 31-0], which recommends the use of the term coefficient when it links two quantities A and B (equation 1) that have different dimensions. The term factor should only be used for k when it is used to link the terms A and B that have the same dimensions A=k.B. However, in a typical calibration, an ion chamber is calibrated in terms of a physical quantity such as air kerma, dose to water, ambient dose equivalent, etc. If the chamber is calibrated together with its electrometer, then the calibration refers to the physical quantity to be measured per electrometer unit reading. In this case, the terms referred have different dimensions. The adoption by the Agency of the term coefficient to express the results of calibrations is consistent with the 'International vocabulary of basic and general terms in metrology' prepared jointly by the BIPM, IEC, ISO, OIML and other organizations. The BIPM has changed from factor to coefficient. The authors believe that this is more than just a matter of semantics and recommend that the SSDL network members adopt this change in terminology. (author)

  12. Calibration of an inertial-magnetic measurement unit without external equipment, in the presence of dynamic magnetic disturbances

    International Nuclear Information System (INIS)

    Metge, J; Giremus, A; Mégret, R; Berthoumieu, Y; Décamps, T

    2014-01-01

    Inertial-magnetic measurement units are inexpensive sensors, widely used in electronic systems (smartphones, GPS, micro-UAV, etc). However the precision of these sensors is highly dependent on their calibration. This article proposes a complete solution to calibrate the sensors (accelerometers, gyrometers and magnetometers), the inter-sensor rotations and the dynamic disturbances of the magnetic field due to the immediate environment. Contrary to most of the existing techniques, the proposed method does not necessitate any external equipment, apart from the sensors already included in the system. The calibration can be performed by hand manipulation by the final user. Simulations and experiments show the advantages of the proposed approach. (paper)

  13. Calibration by Hydrological Response Unit of a National Hydrologic Model to Improve Spatial Representation and Distribution of Parameters

    Science.gov (United States)

    Norton, P. A., II

    2015-12-01

    The U. S. Geological Survey is developing a National Hydrologic Model (NHM) to support consistent hydrologic modeling across the conterminous United States (CONUS). The Precipitation-Runoff Modeling System (PRMS) simulates daily hydrologic and energy processes in watersheds, and is used for the NHM application. For PRMS each watershed is divided into hydrologic response units (HRUs); by default each HRU is assumed to have a uniform hydrologic response. The Geospatial Fabric (GF) is a database containing initial parameter values for input to PRMS and was created for the NHM. The parameter values in the GF were derived from datasets that characterize the physical features of the entire CONUS. The NHM application is composed of more than 100,000 HRUs from the GF. Selected parameter values commonly are adjusted by basin in PRMS using an automated calibration process based on calibration targets, such as streamflow. Providing each HRU with distinct values that captures variability within the CONUS may improve simulation performance of the NHM. During calibration of the NHM by HRU, selected parameter values are adjusted for PRMS based on calibration targets, such as streamflow, snow water equivalent (SWE) and actual evapotranspiration (AET). Simulated SWE, AET, and runoff were compared to value ranges derived from multiple sources (e.g. the Snow Data Assimilation System, the Moderate Resolution Imaging Spectroradiometer (i.e. MODIS) Global Evapotranspiration Project, the Simplified Surface Energy Balance model, and the Monthly Water Balance Model). This provides each HRU with a distinct set of parameter values that captures the variability within the CONUS, leading to improved model performance. We present simulation results from the NHM after preliminary calibration, including the results of basin-level calibration for the NHM using: 1) default initial GF parameter values, and 2) parameter values calibrated by HRU.

  14. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John F. Schabron; Joseph F. Rovani; Susan S. Sorini

    2007-03-31

    The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005, requires that calibration of mercury continuous emissions monitors (CEMs) be performed with NIST-traceable standards. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The traceability protocol will be written by EPA. Traceability will be based on the actual analysis of the output of each calibration unit at several concentration levels ranging from about 2-40 ug/m{sup 3}, and this analysis will be directly traceable to analyses by NIST using isotope dilution inductively coupled plasma/mass spectrometry (ID ICP/MS) through a chain of analyses linking the calibration unit in the power plant to the NIST ID ICP/MS. Prior to this project, NIST did not provide a recommended mercury vapor pressure equation or list mercury vapor pressure in its vapor pressure database. The NIST Physical and Chemical Properties Division in Boulder, Colorado was subcontracted under this project to study the issue in detail and to recommend a mercury vapor pressure equation that the vendors of mercury vapor pressure calibration units can use to calculate the elemental mercury vapor concentration in an equilibrium chamber at a particular temperature. As part of this study, a preliminary evaluation of calibration units from five vendors was made. The work was performed by NIST in Gaithersburg, MD and Joe Rovani from WRI who traveled to NIST as a Visiting Scientist.

  15. Mechanics of log calibration

    International Nuclear Information System (INIS)

    Waller, W.C.; Cram, M.E.; Hall, J.E.

    1975-01-01

    For any measurement to have meaning, it must be related to generally accepted standard units by a valid and specified system of comparison. To calibrate well-logging tools, sensing systems are designed which produce consistent and repeatable indications over the range for which the tool was intended. The basics of calibration theory, procedures, and calibration record presentations are reviewed. Calibrations for induction, electrical, radioactivity, and sonic logging tools will be discussed. The authors' intent is to provide an understanding of the sources of errors, of the way errors are minimized in the calibration process, and of the significance of changes in recorded calibration data

  16. A Review of Sensor Calibration Monitoring for Calibration Interval Extension in Nuclear Power Plants

    Energy Technology Data Exchange (ETDEWEB)

    Coble, Jamie B.; Meyer, Ryan M.; Ramuhalli, Pradeep; Bond, Leonard J.; Hashemian, Hash; Shumaker, Brent; Cummins, Dara

    2012-08-31

    Currently in the United States, periodic sensor recalibration is required for all safety-related sensors, typically occurring at every refueling outage, and it has emerged as a critical path item for shortening outage duration in some plants. Online monitoring can be employed to identify those sensors that require calibration, allowing for calibration of only those sensors that need it. International application of calibration monitoring, such as at the Sizewell B plant in United Kingdom, has shown that sensors may operate for eight years, or longer, within calibration tolerances. This issue is expected to also be important as the United States looks to the next generation of reactor designs (such as small modular reactors and advanced concepts), given the anticipated longer refueling cycles, proposed advanced sensors, and digital instrumentation and control systems. The U.S. Nuclear Regulatory Commission (NRC) accepted the general concept of online monitoring for sensor calibration monitoring in 2000, but no U.S. plants have been granted the necessary license amendment to apply it. This report presents a state-of-the-art assessment of online calibration monitoring in the nuclear power industry, including sensors, calibration practice, and online monitoring algorithms. This assessment identifies key research needs and gaps that prohibit integration of the NRC-approved online calibration monitoring system in the U.S. nuclear industry. Several needs are identified, including the quantification of uncertainty in online calibration assessment; accurate determination of calibration acceptance criteria and quantification of the effect of acceptance criteria variability on system performance; and assessment of the feasibility of using virtual sensor estimates to replace identified faulty sensors in order to extend operation to the next convenient maintenance opportunity. Understanding the degradation of sensors and the impact of this degradation on signals is key to

  17. The energetic NeUtral Atom Detector Unit (NUADU) for China's Double Star Mission and its calibration

    Energy Technology Data Exchange (ETDEWEB)

    McKenna-Lawlor, Susan E-mail: stil@may.ie; Balaz, Jan; Strharsky, Igor; Barabash, Stas; Brinkfeldt, Klas; Li Lu; Shen Chao; Shi Jiankui; Zong Qingang; Kudela, Karel; Fu Suiyan; Roelof, E.C.; Brandt, Pontus C. son; Dandouras, Iannis

    2004-09-11

    An account is provided of an advanced Energetic NeUtral Atom Detector Unit (NUADU) designed for China's Double Star Mission. Special emphasis is given to describing the detector head of the instrument and its calibration.

  18. Preliminary Results of BTDF Calibration of Transmissive Solar Diffusers for Remote Sensing

    Science.gov (United States)

    Georgiev, Georgi T.; Butler, James J.; Thome, Kurt; Cooksey, Catherine; Ding, Leibo

    2016-01-01

    Satellite instruments operating in the reflected solar wavelength region require accurate and precise determination of the optical properties of their diffusers used in pre-flight and post-flight calibrations. The majority of recent and current space instruments use reflective diffusers. As a result, numerous Bidirectional Reflectance Distribution Function (BRDF) calibration comparisons have been conducted between the National Institute of Standards and Technology (NIST) and other industry and university-based metrology laboratories. However, based on literature searches and communications with NIST and other laboratories, no Bidirectional Transmittance Distribution Function (BTDF) measurement comparisons have been conducted between National Measurement Laboratories (NMLs) and other metrology laboratories. On the other hand, there is a growing interest in the use of transmissive diffusers in the calibration of satellite, air-borne, and ground-based remote sensing instruments. Current remote sensing instruments employing transmissive diffusers include the Ozone Mapping and Profiler Suite instrument (OMPS) Limb instrument on the Suomi-National Polar-orbiting Partnership (S-NPP) platform,, the Geostationary Ocean Color Imager (GOCI) on the Korea Aerospace Research Institute's (KARI) Communication, Ocean, and Meteorological Satellite (COMS), the Ozone Monitoring Instrument (OMI) on NASA's Earth Observing System (EOS) Aura platform, the Tropospheric Emissions: Monitoring of Pollution (TEMPO) instrument and the Geostationary Environmental Monitoring Spectrometer (GEMS).. This ensemble of instruments requires validated BTDF measurements of their on-board transmissive diffusers from the ultraviolet through the near infrared. This paper presents the preliminary results of a BTDF comparison between the NASA Diffuser Calibration Laboratory (DCL) and NIST on quartz and thin Spectralon samples.

  19. Preliminary results of BTDF calibration of transmissive solar diffusers for remote sensing

    Science.gov (United States)

    Georgiev, Georgi T.; Butler, James J.; Thome, Kurt; Cooksey, Catherine; Ding, Leibo

    2016-09-01

    Satellite instruments operating in the reflected solar wavelength region require accurate and precise determination of the optical properties of their diffusers used in pre-flight and post-flight calibrations. The majority of recent and current space instruments use reflective diffusers. As a result, numerous Bidirectional Reflectance Distribution Function (BRDF) calibration comparisons have been conducted between the National Institute of Standards and Technology (NIST) and other industry and university-based metrology laboratories. However, based on literature searches and communications with NIST and other laboratories, no Bidirectional Transmittance Distribution Function (BTDF) measurement comparisons have been conducted between National Measurement Laboratories (NMLs) and other metrology laboratories. On the other hand, there is a growing interest in the use of transmissive diffusers in the calibration of satellite, air-borne, and ground-based remote sensing instruments. Current remote sensing instruments employing transmissive diffusers include the Ozone Mapping and Profiler Suite instrument (OMPS) Limb instrument on the Suomi-National Polar-orbiting Partnership (S-NPP) platform,, the Geostationary Ocean Color Imager (GOCI) on the Korea Aerospace Research Institute's (KARI) Communication, Ocean, and Meteorological Satellite (COMS), the Ozone Monitoring Instrument (OMI) on NASA's Earth Observing System (EOS) Aura platform, the Tropospheric Emissions: Monitoring of Pollution (TEMPO) instrument and the Geostationary Environmental Monitoring Spectrometer (GEMS).. This ensemble of instruments requires validated BTDF measurements of their onboard transmissive diffusers from the ultraviolet through the near infrared. This paper presents the preliminary results of a BTDF comparison between the NASA Diffuser Calibration Laboratory (DCL) and NIST on quartz and thin Spectralon samples.

  20. Primary calibration in acoustics metrology

    International Nuclear Information System (INIS)

    Milhomem, T A Bacelar; Soares, Z M Defilippo

    2015-01-01

    SI unit in acoustics is realized by the reciprocity calibrations of laboratory standard microphones in pressure field, free field and diffuse field. Calibrations in pressure field and in free field are already consolidated and the Inmetro already done them. Calibration in diffuse field is not yet consolidated, however, some national metrology institutes, including Inmetro, are conducting researches on this subject. This paper presents the reciprocity calibration, the results of Inmetro in recent key comparisons and the research that is being developed for the implementation of reciprocity calibration in diffuse field

  1. Development of pre-critical excore detector linear subchannel calibration method

    International Nuclear Information System (INIS)

    Choi, Yoo Sun; Goo, Bon Seung; Cha, Kyun Ho; Lee, Chang Seop; Kim, Yong Hee; Ahn, Chul Soo; Kim, Man Soo

    2001-01-01

    The improved pre-critical excore detector linear subchannel calibration method has been developed to improve the applicability of pre-critical calibration method. The existing calibration method does not always guarantee the accuracy of pre-critical calibration because the calibration results of the previous cycle are not reflected into the current cycle calibration. The developed method has a desirable feature that calibration error would not be propagated in the following cycles since the calibration data determined in previous cycle is incorporated in the current cycle calibration. The pre-critical excore detector linear calibration is tested for YGN unit 3 and UCN unit 3 to evaluate its characteristics and accuracy

  2. Preliminary application of a novel algorithm to monitor changes in pre-flight total peripheral resistance for prediction of post-flight orthostatic intolerance in astronauts

    Science.gov (United States)

    Arai, Tatsuya; Lee, Kichang; Stenger, Michael B.; Platts, Steven H.; Meck, Janice V.; Cohen, Richard J.

    2011-04-01

    Orthostatic intolerance (OI) is a significant challenge for astronauts after long-duration spaceflight. Depending on flight duration, 20-80% of astronauts suffer from post-flight OI, which is associated with reduced vascular resistance. This paper introduces a novel algorithm for continuously monitoring changes in total peripheral resistance (TPR) by processing the peripheral arterial blood pressure (ABP). To validate, we applied our novel mathematical algorithm to the pre-flight ABP data previously recorded from twelve astronauts ten days before launch. The TPR changes were calculated by our algorithm and compared with the TPR value estimated using cardiac output/heart rate before and after phenylephrine administration. The astronauts in the post-flight presyncopal group had lower pre-flight TPR changes (1.66 times) than those in the non-presyncopal group (2.15 times). The trend in TPR changes calculated with our algorithm agreed with the TPR trend calculated using measured cardiac output in the previous study. Further data collection and algorithm refinement are needed for pre-flight detection of OI and monitoring of continuous TPR by analysis of peripheral arterial blood pressure.

  3. FTIR Calibration Methods and Issues

    Science.gov (United States)

    Perron, Gaetan

    Over the past 10 years, several space-borne FTIR missions were launched for atmospheric research, environmental monitoring and meteorology. One can think of the Michelson Interferometer for Passive Atmospheric Sounding (MIPAS) launched by the European Space Agency, the Atmospheric Chemistry Experiment (ACE) launched by the Canadian Space Agency, the Tropospheric Emission Spectrometer (TES) launched by NASA and the Infrared Atmospheric Sounding Interferometer (IASI) launched by Eumetsat in Europe. Others are near to be launched, namely the Cross-track Infrared Sounder (CrIS) from the Integrated Program Of- fice in the United States and the Thermal And Near infrared Sensor for carbon Observation (TANSO) from the Japan Aerospace Exploration Agency. Moreover, several missions under definition foresee the use of this technology as sensor, e.g. Meteosat Third Generation (MTG), Eumetsat Polar System (EPS) and the Premier mission, one of the six candidates of the next ESA Earth Explorer Core Mission. In order to produce good quality products, calibration is essential. Calibrated data is the output of three main sub-systems that are tightly coupled: the instrument, the calibration targets and the level 1B processor. Calibration requirements must be carefully defined and propagated to each sub-system. Often, they are carried out by different parties which add to the complexity. Under budget and schedule pressure, some aspects are sometimes neglected and jeopardized final quality. For space-borne FTIR, level 1B outputs are spectra that are radiometrically, spectrally calibrated and geolocated. Radiometric calibration means to assign an intensity value in units to the y-axis. Spectral calibration means to assign to the x-axis the proper frequency value in units. Finally, geolocated means to assign a target position over the earth geoid i.e. longitude, latitude and altitude. This paper will present calibration methods and issues related to space-borne FTIR missions, e.g. two

  4. Radiometric Cross-Calibration of the Chilean Satellite FASat-C Using RapidEye and EO-1 Hyperion Data and a Simultaneous Nadir Overpass Approach

    Directory of Open Access Journals (Sweden)

    Carolina Barrientos

    2016-07-01

    Full Text Available The absolute radiometric calibration of a satellite sensor is the critical factor that ensures the usefulness of the acquired data for quantitative applications on remote sensing. This work presents the results of the first cross-calibration of the sensor on board the Sistema Satelital de Observación de la Tierra (SSOT Chilean satellite or Air Force Satellite FASat-C. RapidEye-MSI was chosen as the reference sensor, and a simultaneous Nadir Overpass Approach (SNO was applied. The biases caused by differences in the spectral responses of both instruments were compensated through an adjustment factor derived from EO-1 Hyperion data. Through this method, the variations affecting the radiometric response of New AstroSat Optical Modular Instrument (NAOMI-1, have been corrected based on collections over the Frenchman Flat calibration site. The results of a preliminary evaluation of the pre-flight and updated coefficients have shown a significant improvement in the accuracy of at-sensor radiances and TOA reflectances: an average agreement of 2.63% (RMSE was achieved for the multispectral bands of both instruments. This research will provide a basis for the continuity of calibration and validation tasks of future Chilean space missions.

  5. Poster Abstract: Automatic Calibration of Device Attitude in Inertial Measurement Unit Based Traffic Probe Vehicles

    KAUST Repository

    Mousa, Mustafa

    2016-04-28

    Probe vehicles consist in mobile traffic sensor networks that evolve with the flow of vehicles, transmitting velocity and position measurements along their path, generated using GPSs. To address the urban positioning issues of GPSs, we propose to replace them with inertial measurement units onboard vehicles, to estimate vehicle location and attitude using inertial data only. While promising, this technology requires one to carefully calibrate the orientation of the device inside the vehicle to be able to process the acceleration and rate gyro data. In this article, we propose a scheme that can perform this calibration automatically by leveraging the kinematic constraints of ground vehicles, and that can be implemented on low-end computational platforms. Preliminary testing shows that the proposed scheme enables one to accurately estimate the actual accelerations and rotation rates in the vehicle coordinates. © 2016 IEEE.

  6. An in situ hand calibration method using a pseudo-observation scheme for low-end inertial measurement units

    International Nuclear Information System (INIS)

    Li, You; Niu, Xiaoji; Zhang, Quan; Zhang, Hongping; Shi, Chuang

    2012-01-01

    MEMS chips have become ideal candidates for various applications since they are small sized, light weight, have low power consumption and are extremely low cost and reliable. However, the performance of MEMS sensors, especially their biases and scale factors, is highly dependent on environmental conditions such as temperature. Thus a quick and convenient calibration is needed to be conducted by users in field without any external equipment or any expert knowledge of calibration. A novel and efficient in situ hand calibration method is presented to meet these demands in this paper. The algorithm of the proposed calibration method makes use of the navigation algorithm of the loosely-coupled GPS/INS integrated systems, but replaces the GPS observations with a kind of pseudo-observations, which can be stated as follows: if an inertial measurement unit (IMU) was rotating approximately around its measurement center, the range of its position and its linear velocity both would be within a limited scope. Using a Kalman filtering algorithm, the biases and scale factors of both accelerometer triad and gyroscope triad can be calibrated together within a short period (about 30 s), requiring only motions by hands. Real test results show that the proposed method is suitable for most consumer grade MEMS IMUs due to its zero cost, easy operation and sufficient accuracy. (paper)

  7. Ratioing methods for in-flight response calibration of space-based spectro-radiometers, operating in the solar spectral region

    Science.gov (United States)

    Lobb, Dan

    2017-11-01

    One of the most significant problems for space-based spectro-radiometer systems, observing Earth from space in the solar spectral band (UV through short-wave IR), is in achievement of the required absolute radiometric accuracy. Classical methods, for example using one or more sun-illuminated diffusers as reflectance standards, do not generally provide methods for monitoring degradation of the in-flight reference after pre-flight characterisation. Ratioing methods have been proposed that provide monitoring of degradation of solar attenuators in flight, thus in principle allowing much higher confidence in absolute response calibration. Two example methods are described. It is shown that systems can be designed for relatively low size and without significant additions to the complexity of flight hardware.

  8. Design and realization of an active SAR calibrator for TerraSAR-X

    Science.gov (United States)

    Dummer, Georg; Lenz, Rainer; Lutz, Benjamin; Kühl, Markus; Müller-Glaser, Klaus D.; Wiesbeck, Werner

    2005-10-01

    TerraSAR-X is a new earth observing satellite which will be launched in spring 2006. It carries a high resolution X-band SAR sensor. For high image data quality, accurate ground calibration targets are necessary. This paper describes a novel system concept for an active and highly integrated, digitally controlled SAR system calibrator. A total of 16 active transponder and receiver systems and 17 receiver only systems will be fabricated for a calibration campaign. The calibration units serve for absolute radiometric calibration of the SAR image data. Additionally, they are equipped with an extra receiver path for two dimensional satellite antenna pattern recognition. The calibrator is controlled by a dedicated digital Electronic Control Unit (ECU). The different voltages needed by the calibrator and the ECU are provided by the third main unit called Power Management Unit (PMU).

  9. Validation of a densimeter calibration procedure for a secondary calibration laboratory

    International Nuclear Information System (INIS)

    Alpizar Herrera, Juan Carlos

    2014-01-01

    A survey was conducted to quantify the need for calibration of a density measurement instrument at the research units at the Sede Rodrigo Facio of the Universidad de Costa Rica. A calibration procedure was documented for the instrument that presented the highest demand in the survey by the calibration service. A study of INTE-ISO/IEC 17025: 2005 and specifically in section 5.4 of this standard was done, to document the procedure for calibrating densimeters. Densimeter calibration procedures and standards were sought from different national and international sources. The method of hydrostatic weighing or Cuckow method was the basis of the defined procedure. Documenting the calibration procedure and creating other documents was performed for data acquisition log, intermediate calculation log and calibration certificate copy. A veracity test was performed using as reference laboratory a laboratory of calibration secondary national as part of the validation process of the documented procedure. The results of the E_n statistic of 0.41; 0.34 and 0.46 for the calibration points 90%, 50% and 10% were obtained for the densimeter scale respectively. A reproducibility analysis of the method was performed with satisfactory results. Different suppliers were contacted to estimate the economic costs of the equipment and materials, needed to develop the documented method of densimeter calibration. The acquisition of an analytical balance was recommended, instead of a precision scale, in order to improve the results obtained with the documented method [es

  10. Technical Note: Calibration device for the krypton hygrometer KH20

    Directory of Open Access Journals (Sweden)

    T. Foken

    2012-08-01

    Full Text Available A calibration device for krypton hygrometers (KH20, Campbell Scientific, Inc. with variable path length is presented. This unit allows for in-situ calibrations of the krypton hygrometer, which is typically not very stable over time, during measuring campaigns. It was constructed mainly for application at high altitudes and low temperatures, where further improvements are needed to the IR-hygrometers which are normally used. The changing path length requires that a changing concentration of the absorber be simulated. Because oxygen absorbs more strongly than water vapour, the calibration is made against oxygen and transferred to water vapour. The design of the calibration instrument is made as one unit containing a stepper motor system, PC and humidity sensor. For the calibration, it is necessary to install the krypton hygrometer on this unit.

  11. The GERDA calibration system

    Energy Technology Data Exchange (ETDEWEB)

    Baudis, Laura; Froborg, Francis; Tarka, Michael; Bruch, Tobias; Ferella, Alfredo [Physik-Institut, Universitaet Zuerich (Switzerland); Collaboration: GERDA-Collaboration

    2012-07-01

    A system with three identical custom made units is used for the energy calibration of the GERDA Ge diodes. To perform a calibration the {sup 228}Th sources are lowered from the parking positions at the top of the cryostat. Their positions are measured by two independent modules. One, the incremental encoder, counts the holes in the perforated steel band holding the sources, the other measures the drive shaft's angular position even if not powered. The system can be controlled remotely by a Labview program. The calibration data is analyzed by an iterative calibration algorithm determining the calibration functions for different energy reconstruction algorithms and the resolution of several peaks in the {sup 228}Th spectrum is determined. A Monte Carlo simulation using the GERDA simulation software MAGE has been performed to determine the background induced by the sources in the parking positions.

  12. A novel method of calibrating a MEMS inertial reference unit on a turntable under limited working conditions

    Science.gov (United States)

    Lu, Jiazhen; Liang, Shufang; Yang, Yanqiang

    2017-10-01

    Micro-electro-mechanical systems (MEMS) inertial measurement devices tend to be widely used in inertial navigation systems and have quickly emerged on the market due to their characteristics of low cost, high reliability and small size. Calibration is the most effective way to remove the deterministic error of an inertial reference unit (IRU), which in this paper consists of three orthogonally mounted MEMS gyros. However, common testing methods in the lab cannot predict the corresponding errors precisely when the turntable’s working condition is restricted. In this paper, the turntable can only provide a relatively small rotation angle. Moreover, the errors must be compensated exactly because of the great effect caused by the high angular velocity of the craft. To deal with this question, a new method is proposed to evaluate the MEMS IRU’s performance. In the calibration procedure, a one-axis table that can rotate a limited angle in the form of a sine function is utilized to provide the MEMS IRU’s angular velocity. A new algorithm based on Fourier series is designed to calculate the misalignment and scale factor errors. The proposed method is tested in a set of experiments, and the calibration results are compared to a traditional calibration method performed under normal working conditions to verify their correctness. In addition, a verification test in the given rotation speed is implemented for further demonstration.

  13. Quality control of radiopharmaceutical dose calibrators in nuclear medicine unit

    International Nuclear Information System (INIS)

    Oliveira, C.F.M.; Lucindo Junior, C.R.; Lopes Filho, F.J.

    2015-01-01

    As part of the program to ensure quality in nuclear medicine unit, in addition to diagnostic procedures, are evaluated activity meters, which is intended to measure the aliquot of radiation of radionuclides and / or radiopharmaceuticals that are administered to patients undergoing diagnostic investigation and / or therapeutic treatment. The good operating condition of dose calibrators is essential to ensure efficiency, safety and reliability of the measurements, once the lack of accuracy in the responses of these equipments can cause significant errors in the activity administered to the patient and may result in poor quality images resulting in the repetition of examis and interference in the successful treatment of the patient. This study aims to, considering the need for constant evaluation of the functioning of the activity meters and the fact that this issue be part the responsibilities of the professional of radiology, perform quality control testing of these instruments in relation to the most recent norm of National Commission of nuclear Energy (CNEN-NN 3:05) in Brazil, that is also in according to the international standards and reference values established during acceptance testing of these instruments in a nuclear medicine service. For this, was made a review of specific literature and the use of barium, cobalt and cesium to the tests in a nuclear medicine service of the state of Pernambuco in Brazil. The obtained results of the specific tests utilized to verify the correct working of the dose calibrators show coherency with the resolutions of the CNEN-NN 3:05 and are also in agreement with the international standards to that the measurement of activities be made with accurate results and thereby contribute to the proper functioning of nuclear medicine service. (authors)

  14. Establishing BRDF calibration capabilities through shortwave infrared

    Science.gov (United States)

    Georgiev, Georgi T.; Butler, James J.; Thome, Kurt; Cooksey, Catherine; Ding, Leibo

    2017-09-01

    Satellite instruments operating in the reflective solar wavelength region require accurate and precise determination of the Bidirectional Reflectance Distribution Functions (BRDFs) of the laboratory and flight diffusers used in their pre-flight and on-orbit calibrations. This paper advances that initial work and presents a comparison of spectral Bidirectional Reflectance Distribution Function (BRDF) and Directional Hemispherical Reflectance (DHR) of Spectralon*, a common material for laboratory and onorbit flight diffusers. A new measurement setup for BRDF measurements from 900 nm to 2500 nm located at NASA Goddard Space Flight Center (GSFC) is described. The GSFC setup employs an extended indium gallium arsenide detector, bandpass filters, and a supercontinuum light source. Comparisons of the GSFC BRDF measurements in the shortwave infrared (SWIR) with those made by the National Institute of Standards and Technology (NIST) Spectral Tri-function Automated Reference Reflectometer (STARR) are presented. The Spectralon sample used in this study was 2 inch diameter, 99% white pressed and sintered Polytetrafluoroethylene (PTFE) target. The NASA/NIST BRDF comparison measurements were made at an incident angle of 0° and viewing angle of 45° . Additional BRDF data not compared to NIST were measured at additional incident and viewing angle geometries and are not presented here. The total combined uncertainty for the measurement of BRDF in the SWIR range made by the GSFC scatterometer is less than 1% (k = 1). This study is in support of the calibration of the Radiation Budget Instrument (RBI) and Visible Infrared Imaging Radiometer Suit (VIIRS) instruments of the Joint Polar Satellite System (JPSS) and other current and future NASA remote sensing missions operating across the reflected solar wavelength region.

  15. Radioactivity measurement of 18F in 16 ml vials for calibration of radionuclide calibrators

    International Nuclear Information System (INIS)

    Wurdiyanto, Gatot; Marsoem, Pujadi; Candra, Hermawan; Wijono, Paidi

    2012-01-01

    Fluorine-18 is obtained through the reaction 18 O(p, n) 18 F using a cyclotron that is situated in a hospital in Jakarta. Standardization of the 18 F solution is performed by gamma spectrometry using calibration sources of 152 Eu, 60 Co and 137 Cs that have traceability to the International System of units (SI). The activities in the 16 ml vials that were used for calibrating the radionuclide calibrators were between 1 and 2 GBq, with expanded uncertainties of 3.8%. The expanded uncertainty, at a coverage factor of k=2, on the derived calibration factor for the radionuclide calibrator was 6.6%. - Highlights: ► PTKMR–BATAN as a NMI of Indonesia is required to have procedures to calibrate the radionuclide calibrators. ► Standardizations were carried out on a solution of [ 18 F]FDG using gamma spectrometry. ► The volume of 18 F solutions used was 16 ml because this is the volume often used in hospitals. ► The Secondary Standard ionization chamber is a CRC-7BT Capintec radionuclide calibrator. ► A dial setting for 16 ml of [ 18 F]FDG solution in a vial is 443 for the Capintec dose calibrator.

  16. Automated Attitude Sensor Calibration: Progress and Plans

    Science.gov (United States)

    Sedlak, Joseph; Hashmall, Joseph

    2004-01-01

    This paper describes ongoing work a NASA/Goddard Space Flight Center to improve the quality of spacecraft attitude sensor calibration and reduce costs by automating parts of the calibration process. The new calibration software can autonomously preview data quality over a given time span, select a subset of the data for processing, perform the requested calibration, and output a report. This level of automation is currently being implemented for two specific applications: inertial reference unit (IRU) calibration and sensor alignment calibration. The IRU calibration utility makes use of a sequential version of the Davenport algorithm. This utility has been successfully tested with simulated and actual flight data. The alignment calibration is still in the early testing stage. Both utilities will be incorporated into the institutional attitude ground support system.

  17. Absolute radiometric calibration of Landsat using a pseudo invariant calibration site

    Science.gov (United States)

    Helder, D.; Thome, K.J.; Mishra, N.; Chander, G.; Xiong, Xiaoxiong; Angal, A.; Choi, Tae-young

    2013-01-01

    Pseudo invariant calibration sites (PICS) have been used for on-orbit radiometric trending of optical satellite systems for more than 15 years. This approach to vicarious calibration has demonstrated a high degree of reliability and repeatability at the level of 1-3% depending on the site, spectral channel, and imaging geometries. A variety of sensors have used this approach for trending because it is broadly applicable and easy to implement. Models to describe the surface reflectance properties, as well as the intervening atmosphere have also been developed to improve the precision of the method. However, one limiting factor of using PICS is that an absolute calibration capability has not yet been fully developed. Because of this, PICS are primarily limited to providing only long term trending information for individual sensors or cross-calibration opportunities between two sensors. This paper builds an argument that PICS can be used more extensively for absolute calibration. To illustrate this, a simple empirical model is developed for the well-known Libya 4 PICS based on observations by Terra MODIS and EO-1 Hyperion. The model is validated by comparing model predicted top-of-atmosphere reflectance values to actual measurements made by the Landsat ETM+ sensor reflective bands. Following this, an outline is presented to develop a more comprehensive and accurate PICS absolute calibration model that can be Système international d'unités (SI) traceable. These initial concepts suggest that absolute calibration using PICS is possible on a broad scale and can lead to improved on-orbit calibration capabilities for optical satellite sensors.

  18. Pre-flight physical simulation test of HIMES reentry test vehicle

    Science.gov (United States)

    Kawaguchi, Jun'ichiro; Inatani, Yoshifumi; Yonemoto, Koichi; Hosokawa, Shigeru

    ISAS is now developing a small reentry test vehicle, which is 2m long with a 1.5m wing span and weighs about 170 kg, for the purpose of exploring high angle-of-attack aerodynamic attitude control issue in supersonic and hypersonic speed. The flight test, employing 'Rockoon' launch system, is planned as a preliminary design verification for a fully reusable winged rocket named HIMES (Highly Maneuverable Experimental Space) vehicle. This paper describes the results of preflight ground test using a motion table system. This ground system test is called 'physical simulation' aimed at: (1) functional verification of side-jet system, aerodynamic surface actuators, battery and onboard avionics; and (2) guidance and control law evaluation, in total hardware-in-the-loop system. The pressure of side-jet nozzles was measured to provide exact thrust characteristics of reaction control. The dynamics of vehicle motion was calculated in real-time by the ground simulation computer.

  19. Physiotherapy ultrasound calibrations

    International Nuclear Information System (INIS)

    Gledhill, M.

    1996-01-01

    Calibration of physiotherapy ultrasound equipment has long been a problem. Numerous surveys around the world over the past 20 years have all found that only a low percentage of the units tested had an output within 30% of that indicatd. In New Zealand, a survey carried out by the NRL in 1985 found that only 24% had an output, at the maximum setting, within + or - 20% of that indicated. The present performance Standard for new equipment (NZS 3200.2.5:1992) requires that the measured output should not deviate from that indicated by more than + or - 30 %. This may be tightened to + or - 20% in the next few years. Any calibration is only as good as the calibration equipment. Some force balances can be tested with small weights to simulate the force exerted by an ultrasound beam, but with others this is not possible. For such balances, testing may only be feasible with a calibrated source which could be used like a transfer standard. (author). 4 refs., 3 figs

  20. Overview of Pre-Flight Physical Training, In-Flight Exercise Countermeasures and the Post-Flight Reconditioning Program for International Space Station Astronauts

    Science.gov (United States)

    Kerstman, Eric

    2011-01-01

    International Space Station (ISS) astronauts receive supervised physical training pre-flight, utilize exercise countermeasures in-flight, and participate in a structured reconditioning program post-flight. Despite recent advances in exercise hardware and prescribed exercise countermeasures, ISS crewmembers are still found to have variable levels of deconditioning post-flight. This presentation provides an overview of the astronaut medical certification requirements, pre-flight physical training, in-flight exercise countermeasures, and the post-flight reconditioning program. Astronauts must meet medical certification requirements on selection, annually, and prior to ISS missions. In addition, extensive physical fitness testing and standardized medical assessments are performed on long duration crewmembers pre-flight. Limited physical fitness assessments and medical examinations are performed in-flight to develop exercise countermeasure prescriptions, ensure that the crewmembers are physically capable of performing mission tasks, and monitor astronaut health. Upon mission completion, long duration astronauts must re-adapt to the 1 G environment, and be certified as fit to return to space flight training and active duty. A structured, supervised postflight reconditioning program has been developed to prevent injuries, facilitate re-adaptation to the 1 G environment, and subsequently return astronauts to training and space flight. The NASA reconditioning program is implemented by the Astronaut Strength, Conditioning, and Rehabilitation (ASCR) team and supervised by NASA flight surgeons. This program has evolved over the past 10 years of the International Space Station (ISS) program and has been successful in ensuring that long duration astronauts safely re-adapt to the 1 g environment and return to active duty. Lessons learned from this approach to managing deconditioning can be applied to terrestrial medicine and future exploration space flight missions.

  1. Calibration of whole-body counters for transuranic radionuclides by using total-body donations to the United States Transuranium Registry

    International Nuclear Information System (INIS)

    Palmer, H.E.; Heid, K.R.; Breitenstein, B.D.; Durbin, P.W.; McInroy, J.F.

    1985-01-01

    One of the many benefits of the whole-body donation programme of the United States Transuranium Registry is the calibration of whole-body counters by measuring a donated body containing radioactivity before radiochemical analysis of the body occurs. These measurements have provided the best existing calibration factors for measuring 241 Am and 239 Pu in the skeleton of living people. In one case where the body contained an easily measurable quantity of 241 Am, the 241 Am content was found to be the same for identical bones on the right and left sides of the body. The bones from the right side of the body and every other vertebrae were radiochemically analysed and the bones from the left side were reserved. These latter have a known quantity of naturally distributed 241 Am based on the analysis of bones from the right side. These reserved bones are being incorporated into tissue equivalent body parts for future calibrations and interlaboratory comparisons. The measurements on the total body and the reserved bones have already been very useful in determining the contribution of skeletal activity to that of lung measurements and determining a calibration factor for measuring 239 Pu deposited in the skeleton. In the future it is planned to periodically make in vivo measurements on those people participating in the total-body donation programme so that calibration factors can be obtained for lung and liver content after subsequent analysis of these tissues. (author)

  2. ACCESS: integration and pre-flight performance

    Science.gov (United States)

    Kaiser, Mary Elizabeth; Morris, Matthew J.; Aldoroty, Lauren N.; Pelton, Russell; Kurucz, Robert; Peacock, Grant O.; Hansen, Jason; McCandliss, Stephan R.; Rauscher, Bernard J.; Kimble, Randy A.; Kruk, Jeffrey W.; Wright, Edward L.; Orndorff, Joseph D.; Feldman, Paul D.; Moos, H. Warren; Riess, Adam G.; Gardner, Jonathan P.; Bohlin, Ralph; Deustua, Susana E.; Dixon, W. V.; Sahnow, David J.; Perlmutter, Saul

    2017-09-01

    Establishing improved spectrophotometric standards is important for a broad range of missions and is relevant to many astrophysical problems. ACCESS, "Absolute Color Calibration Experiment for Standard Stars", is a series of rocket-borne sub-orbital missions and ground-based experiments designed to enable improvements in the precision of the astrophysical flux scale through the transfer of absolute laboratory detector standards from the National Institute of Standards and Technology (NIST) to a network of stellar standards with a calibration accuracy of 1% and a spectral resolving power of 500 across the 0.35 - 1.7μm bandpass. This paper describes the sub-system testing, payload integration, avionics operations, and data transfer for the ACCESS instrument.

  3. Calibration device for wide range monitor

    International Nuclear Information System (INIS)

    Kodoku, Masaya; Sato, Toshifumi.

    1989-01-01

    The calibration device for a wide range monitor according to the present invention can continuously calibrate the entire counting regions of a wide range monitor. The wide range monitor detect the reactor power in the neutron source region by means of a pulse counting method and detects the reactor power in the intermediate region by means of a cambell method. A calibration signal outputting means is disposed for continuously outputting, as such calibration signals, pulse number varying signals in which the number of pulses per unit time varies depending on the reactor power in the neutron source region to be simulated and amplitude square means varying signal in which the mean square value of amplitude varies depending on the reactor power in the intermediate region to be simulated. By using both of the calibration signals, calibration can be conducted for the nuclear reactor power in the neutron source region and the intermediate region even if the calibration is made over two regions, further, calibration for the period present over the two region can be conducted easily as well. (I.S.)

  4. Intercomparison programme of dose calibration used in nuclear medicine center in Malaysia

    International Nuclear Information System (INIS)

    Norhayati Abdullah; Abdul Aziz Mohd Ramli; Muhammad Jamal Md Isa; Siti Sara Deraman; Shahrul Azlan Azizan; Nor Azlin Azraai; Md Khairusalih Md Zin

    2010-01-01

    Calibration of dose calibrator is significant in order to ensure that the equipment operates optimally and provides accurate and reliable measurements of the total activity of radiopharmaceuticals before being administered into the patients. Through this work, the response between the secondary standard dose calibrator and users radioactivity measurement are obtained by using standard sources such as 57 Co, 133 Ba, 1 '3 7 Cs and 60 Co. The calibration procedure is in accordance with the NPLs (National Physical Laboratory, United Kingdom) document; Guide No. 93[1] and the IAEA (International Atomic Energy Agency) Technical Report Series No. 454 [2] is used as a reference for maintaining secondary standard dose calibrator. A total of 21 units of dose calibrator from eight nuclear medicine departments comprising five hospitals, two medical centres and one production laboratory were calibrated. The measurement results were inter compared with the national standard equipment and a baseline data was established for future comparison as well as dose optimization purposes. Results showed that the overall response of all dose calibrators are within NPLs tolerance limit of ±10 % except for 5 units which exceed the tolerance limit for radionuclide 133 Ba and 57 Co.(author)

  5. CVD calibration light systems specifications. Rev. 0

    International Nuclear Information System (INIS)

    Mcllwain, A. K.

    1992-04-01

    Two prototype Cerenkov Viewing Device Calibration Light systems for the Mark IV CVD have been fabricated. They consist of a maintenance unit that will be used by the IAEA maintenance staff and a field unit that will be used by IAEA inspectors. More detailed information on the design of the calibration units can be obtained from the document SSP-39 and additional information on the Mark IV CVD can be obtained from the operating manual published as Canadian Safeguards Support Program document CSSP 6. The specifications refer to the prototype units which will be demonstrated to the IAEA in 1992 May. Based upon the feedback from the IAEA, the instruments will be changed in the final production models to provide devices that more closely satisfy the needs of the end users

  6. Results of the Phoenix Relative Humidity Sensor Recalibration

    Science.gov (United States)

    Martinez, G.; Fischer, E.; Renno, N. O.

    2017-12-01

    We show results of the recalibration of the Thermal and Electrical Conductivity Probe (TECP) relative humidity (RH) sensor of the Phoenix Mars lander [Zent et al., 2009]. Due to uncertainties in its pre-flight calibration, which partially overlapped the environmental conditions found at the Phoenix landing site [Tamppari et al., 2010], only the raw, unprocessed output of the TECP RH sensor is available in NASA's Planetary Data System (PDS). The sensor's calibration was revised in 2016 to correct for inaccuracies at the lowest temperatures [Zent et al., 2016], but the new processed RH values were not posted in the PDS. We have been using a spare engineering unit of the TECP to recalibrate the sensor in the full range of Phoenix landing site conditions in the Michigan Mars Environmental Chamber (MMEC) [Fischer et al., 2016]. We compare raw output data of the engineering unit in the MMEC with that of the flight unit from the preflight calibration. We observed that the engineering unit's RH sensor output was shifted to higher values compared to the flight unit's output at the same conditions of temperature and humidity. Based on this shift, we use a translation function that fits the in-situ measurements of the flight unit into the engineering unit output space. To improve the accuracy of this function, we use additional observations corresponding to saturated conditions when near-surface fog was observed [Whiteway et al., 2009], as well as observations around noon when the RH is expected to be below 5%. The entire range of conditions observed on the Martian surface is covered in our recalibration. The raw output of the sensor is used to obtain a new calibration function. This allows us to obtain high-level RH data at Martian polar conditions. The recalibrated data will be posted in the PDS. REFERENCES: Fischer, E., et al. (2016), Astrobiology, 16, 12, doi: 10.1089/ast.2016.1525. Tamppari, L. K., et al. (2010), J. Geophys. Res., 115, E00E17, doi:10.1029/2009JE003415

  7. STS payloads mission control study. Volume 2-A, Task 1: Joint products and functions for preflight planning of flight operations, training and simulations

    Science.gov (United States)

    1976-01-01

    Specific products and functions, and associated facility availability, applicable to preflight planning of flight operations were studied. Training and simulation activities involving joint participation of STS and payload operations organizations, are defined. The prelaunch activities required to prepare for the payload flight operations are emphasized.

  8. Field calibration of optical sensors for measuring suspended sediment concentration in the western Mediterranean

    Directory of Open Access Journals (Sweden)

    J. Guillén

    2000-12-01

    Full Text Available The water turbidity measured with optical methods (transmittance and backscattering is usually expressed as beam attenuation coefficient (BAC or formazin turbidity units (FTU. The transformation of these units to volumetric suspended sediment concentration (SSC units is not straightforward, and accurate calibrations are required in order to obtain valuable information on suspended sediment distributions and fluxes. In this paper, data from field calibrations between BAC, FTU and SSC are presented and best-fit calibration curves are shown. These calibrations represent an average from different marine environments of the western Mediterranean (from estuary to continental slope. However, the general curves can only be applied for descriptive or semi-quantitative purposes. Comparison of turbidity measurements using the same sensor with different calibration ranges shows the advantage of simultaneously combining two instruments calibrated in different ranges when significant changes in suspended sediment concentrations are expected.

  9. Calibration assembly for nuclear reactor vessel inspection apparatus

    International Nuclear Information System (INIS)

    Elsner, H.J.

    1981-01-01

    A removable calibration assembly can be utilized to verify the angular mounting of transducers in an array employed in an ultrasonic inpsection apparatus, to calibrate one axis of movement of the array with reference to a starting point, or to measure and calibrate the speed per unit of distance of the transducer's ultrasonic beam in the operating medium. The calibration assembly includes both a large and small reflecting surface separated by known distances, and several large cones, the tips of which are machined or adjusted to angles at which certain of the transducers are to be mounted. Clamping means for securing the calibration assembly to the inspection apparatus at a predetermined orientation is provided

  10. A water flow calorimeter calibration system

    International Nuclear Information System (INIS)

    Ullrich, F.T.

    1983-01-01

    Neutral beam systems are instrumented by several water flow calorimeter systems, and some means is needed to verify the accuracy of such systems and diagnose their failures. This report describes a calibration system for these calorimeters. The calibrator consists of two 24 kilowatt circulation water heaters, with associated controls and instrumentation. The unit can supply power from 0 to 48 kW in five coarse steps and one fine range. Energy is controlled by varying the power and the time of operation of the heaters. The power is measured by means of precision power transducers, and the energy is measured by integrating the power with respect to time. The accuracy of the energy measurement is better than 0.5% when the power supplied is near full scale, and the energy resolution is better than 1 kilojoule. The maximum energy delivered is approximately 50 megajoules. The calorimetry loop to be calibrated is opened, and the calibrator is put in series with the calorimeter heat source. The calorimeter is then operated in its normal fashion, with the calibrator used as the heat source. The calibrator can also be used in a stand alone mode to calibrate calorimeter sensors removed from systems

  11. Phase calibration strategies for synchrotron RF signals

    Energy Technology Data Exchange (ETDEWEB)

    Andreev, Aleksandr [TEMF, Technische Universitaet Darmstadt (Germany); Klingbeil, Harald [TEMF, Technische Universitaet Darmstadt (Germany); GSI Helmholtzzentrum fuer Schwerionenforschung GmbH, Darmstadt (Germany); Lens, Dieter [GSI Helmholtzzentrum fuer Schwerionenforschung GmbH, Darmstadt (Germany)

    2016-07-01

    For the FAIR facility that is currently under construction, the beam quality requirements impose several demands on the low-level RF (LLRF) systems. For example the phase error of the gap voltage of a specific RF cavity must be less than 1 . The RF reference signals for the FAIR synchrotron RF cavity systems are generated by direct digital synthesis modules (DDS) mounted in one crate called Group-DDS. In order to allow performing various multi-harmonic operations, each DDS unit operates at a certain mode defined by the harmonic number that can be changed during the operation. Since the DDS modules generate reference RF signals for different LLRF systems, the precise calibration of units to compensate the different phase response is of importance. The currently used calibration procedure is done with a fixed harmonic number for each module and uses the DDS module configured to the highest harmonic number as a reference. If the harmonic number of the DDS module is changed, one then has to repeat the calibration for the new values. Therefore, a new calibration method with respect to the absolute phases of DDS modules is under development and will be presented.

  12. Unexplored Indoors method for pyranometers calibration traceable to SI

    International Nuclear Information System (INIS)

    Castillo-Matadamas, H A; Molina-Vazquez, J C; Quintero-Torres, R

    2015-01-01

    A method to calibrate pyranometers with direct traceability to the International System of Units (SI) is presented, the method use an electrically calibrated pyroelectric detector (ECPR) as standard and offers numerous advantages over outdoors conventional calibration methods, such as reducing the uncertainty from the reference standard and the final uncertainty of the sensitivity coefficient of the calibrated pyranometer; the measurement uncertainty achieved with this method at normal irradiance is 2.1% for a coverage factor k = 2 and could be reduce if one reduces the uncertainty level of the reference standard

  13. State of the art: two-phase flow calibration techniques

    International Nuclear Information System (INIS)

    Stanley, M.L.

    1977-01-01

    The nuclear community faces a particularly difficult problem relating to the calibration of instrumentation in a two-phase flow steam/water environment. The rationale of the approach to water reactor safety questions in the United States demands that accurate measurements of mass flows in a decompressing two-phase flow be made. An accurate measurement dictates an accurate calibration. This paper addresses three questions relating to the state of the art in two-phase calibration: (1) What do we mean by calibration. (2) What is done now. (3) What should be done

  14. Galileo spacecraft inertial sensors in-flight calibration design

    Science.gov (United States)

    Jahanshahi, M. H.; Lai, J. Y.

    1983-01-01

    The successful navigation of Galileo depends on accurate trajectory correction maneuvers (TCM's) performed during the mission. A set of Inertial Sensor (INS) units, comprised of gyros and accelerometers, mounted on the spacecraft, are utilized to control and monitor the performance of the TCM's. To provide the optimum performance, in-flight calibrations of INS are planned. These calibrations will take place on a regular basis. In this paper, a mathematical description is given of the data reduction technique used in analyzing a typical set of calibration data. The design of the calibration and the inertial sensor error models, necessary for the above analysis, are delineated in detail.

  15. 40 CFR 90.424 - Dilute sampling procedures-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... calibration. 90.424 Section 90.424 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR... “true” value) for the CVS calibration. (Note: In no case should an upstream screen or other restriction... Parameter Symbol Units Sensor-readout tolerances Barometric pressure (corrected) PB kPa ±.340 kPa. Ambient...

  16. Accurate calibration of waveform data measured by the Plasma Wave Experiment on board the ARASE satellite

    Science.gov (United States)

    Kitahara, M.; Katoh, Y.; Hikishima, M.; Kasahara, Y.; Matsuda, S.; Kojima, H.; Ozaki, M.; Yagitani, S.

    2017-12-01

    The Plasma Wave Experiment (PWE) is installed on board the ARASE satellite to measure the electric field in the frequency range from DC to 10 MHz, and the magnetic field in the frequency range from a few Hz to 100 kHz using two dipole wire-probe antennas (WPT) and three magnetic search coils (MSC), respectively. In particular, the Waveform Capture (WFC), one of the receivers of the PWE, can detect electromagnetic field waveform in the frequency range from a few Hz to 20 kHz. The Software-type Wave Particle Interaction Analyzer (S-WPIA) is installed on the ARASE satellite to measure the energy exchange between plasma waves and particles. Since S-WPIA uses the waveform data measured by WFC to calculate the relative phase angle between the wave magnetic field and velocity of energetic electrons, the high-accuracy is required to calibration of both amplitude and phase of the waveform data. Generally, the calibration procedure of the signal passed through a receiver consists of three steps; the transformation into spectra, the calibration by the transfer function of a receiver, and the inverse transformation of the calibrated spectra into the time domain. Practically, in order to reduce the side robe effect, a raw data is filtered by a window function in the time domain before applying Fourier transform. However, for the case that a first order differential coefficient of the phase transfer function of the system is not negligible, the phase of the window function convoluted into the calibrated spectra is shifted differently at each frequency, resulting in a discontinuity in the time domain of the calibrated waveform data. To eliminate the effect of the phase shift of a window function, we suggest several methods to calibrate a waveform data accurately and carry out simulations assuming simple sinusoidal waves as an input signal and using transfer functions of WPT, MSC, and WFC obtained in pre-flight tests. In consequence, we conclude that the following two methods can

  17. Calibration and application of a HPGe gamma spectrometer for in-situ measurements

    International Nuclear Information System (INIS)

    Xiao Xuefu; Yue Qingyu

    1992-02-01

    The principle and methods of the calibration for an in-situ γ spectrometer are introduced. The calibration for a portable HPGe γ spectrometer has been completed. The N f /A(peak count rate per unit activity in soil) and N f /D(peak count rate per unit absorbed dose rate in the air) are listed. The uncertainties of the calibration factors are estimated. The in-situ measurements have been carried out in surroundings near the nuclear facilities and the data are compared with those measured by other methods

  18. Compact optical technique for streak camera calibration

    International Nuclear Information System (INIS)

    Bell, Perry; Griffith, Roger; Hagans, Karla; Lerche, Richard; Allen, Curt; Davies, Terence; Janson, Frans; Justin, Ronald; Marshall, Bruce; Sweningsen, Oliver

    2004-01-01

    To produce accurate data from optical streak cameras requires accurate temporal calibration sources. We have reproduced an older technology for generating optical timing marks that had been lost due to component availability. Many improvements have been made which allow the modern units to service a much larger need. Optical calibrators are now available that produce optical pulse trains of 780 nm wavelength light at frequencies ranging from 0.1 to 10 GHz, with individual pulse widths of approximately 25 ps full width half maximum. Future plans include the development of single units that produce multiple frequencies to cover a wide temporal range, and that are fully controllable via an RS232 interface

  19. Compact optical technique for streak camera calibration

    Science.gov (United States)

    Bell, Perry; Griffith, Roger; Hagans, Karla; Lerche, Richard; Allen, Curt; Davies, Terence; Janson, Frans; Justin, Ronald; Marshall, Bruce; Sweningsen, Oliver

    2004-10-01

    To produce accurate data from optical streak cameras requires accurate temporal calibration sources. We have reproduced an older technology for generating optical timing marks that had been lost due to component availability. Many improvements have been made which allow the modern units to service a much larger need. Optical calibrators are now available that produce optical pulse trains of 780 nm wavelength light at frequencies ranging from 0.1 to 10 GHz, with individual pulse widths of approximately 25 ps full width half maximum. Future plans include the development of single units that produce multiple frequencies to cover a wide temporal range, and that are fully controllable via an RS232 interface.

  20. Inertial Measurement Units-Based Probe Vehicles: Automatic Calibration, Trajectory Estimation, and Context Detection

    KAUST Repository

    Mousa, Mustafa

    2017-12-06

    Most probe vehicle data is generated using satellite navigation systems, such as the Global Positioning System (GPS), Globalnaya navigatsionnaya sputnikovaya Sistema (GLONASS), or Galileo systems. However, because of their high cost, relatively high position uncertainty in cities, and low sampling rate, a large quantity of satellite positioning data is required to estimate traffic conditions accurately. To address this issue, we introduce a new type of traffic monitoring system based on inexpensive inertial measurement units (IMUs) as probe sensors. IMUs as traffic probes pose unique challenges in that they need to be precisely calibrated, do not generate absolute position measurements, and their position estimates are subject to accumulating errors. In this paper, we address each of these challenges and demonstrate that the IMUs can reliably be used as traffic probes. After discussing the sensing technique, we present an implementation of this system using a custom-designed hardware platform, and validate the system with experimental data.

  1. Inertial Measurement Units-Based Probe Vehicles: Automatic Calibration, Trajectory Estimation, and Context Detection

    KAUST Repository

    Mousa, Mustafa; Sharma, Kapil; Claudel, Christian G.

    2017-01-01

    Most probe vehicle data is generated using satellite navigation systems, such as the Global Positioning System (GPS), Globalnaya navigatsionnaya sputnikovaya Sistema (GLONASS), or Galileo systems. However, because of their high cost, relatively high position uncertainty in cities, and low sampling rate, a large quantity of satellite positioning data is required to estimate traffic conditions accurately. To address this issue, we introduce a new type of traffic monitoring system based on inexpensive inertial measurement units (IMUs) as probe sensors. IMUs as traffic probes pose unique challenges in that they need to be precisely calibrated, do not generate absolute position measurements, and their position estimates are subject to accumulating errors. In this paper, we address each of these challenges and demonstrate that the IMUs can reliably be used as traffic probes. After discussing the sensing technique, we present an implementation of this system using a custom-designed hardware platform, and validate the system with experimental data.

  2. Software System for the Calibration of X-Ray Measuring Instruments

    International Nuclear Information System (INIS)

    Gaytan-Gallardo, E.; Tovar-Munoz, V. M.; Cruz-Estrada, P.; Vergara-Martinez, F. J.; Rivero-Gutierrez, T.

    2006-01-01

    A software system that facilities the calibration of X-ray measuring instruments used in medical applications is presented. The Secondary Standard Dosimetry Laboratory (SSDL) of the Nuclear Research National Institute in Mexico (ININ in Spanish), supports activities concerning with ionizing radiations in medical area. One of these activities is the calibration of X-ray measuring instruments, in terms of air kerma or exposure by substitution method in an X-ray beam at a point where the rate has been determined by means of a standard ionization chamber. To automatize this process, a software system has been developed, the calibration system is composed by an X-ray unit, a Dynalizer IIIU X-ray meter by RADCAL, a commercial data acquisition card, the software system and the units to be tested and calibrated. A quality control plan has been applied in the development of the software system, ensuring that quality assurance procedures and standards are being followed

  3. Ultrasonic calibration assembly

    International Nuclear Information System (INIS)

    1981-01-01

    Ultrasonic transducers for in-service inspection of nuclear reactor vessels have several problems associated with them which this invention seeks to overcome. The first is that of calibration or referencing a zero start point for the vertical axis of transducer movement to locate a weld defect. The second is that of verifying the positioning (vertically or at a predetermined angle). Thirdly there is the problem of ascertaining the speed per unit distance in the operating medium of the transducer beam prior to the actual inspection. The apparatus described is a calibration assembly which includes a fixed, generally spherical body having a surface for reflecting an ultrasonic beam from one of the transducers which can be moved until the reflection from the spherical body is the highest amplitude return signal indicating radial alignment from the body. (U.K.)

  4. Preflight screening techniques for centrifuge-simulated suborbital spaceflight.

    Science.gov (United States)

    Pattarini, James M; Blue, Rebecca S; Castleberry, Tarah L; Vanderploeg, James M

    2014-12-01

    Historically, space has been the venue of the healthy individual. With the advent of commercial spaceflight, we face the novel prospect of routinely exposing spaceflight participants (SPFs) with multiple comorbidities to the space environment. Preflight screening procedures must be developed to identify those individuals at increased risk during flight. We examined the responses of volunteers to centrifuge accelerations mimicking commercial suborbital spaceflight profiles to evaluate how potential SFPs might tolerate such forces. We evaluated our screening process for medical approval of subjects for centrifuge participation for applicability to commercial spaceflight operations. All registered subjects completed a medical questionnaire, physical examination, and electrocardiogram. Subjects with identified concerns including cardiopulmonary disease, hypertension, and diabetes were required to provide documentation of their conditions. There were 335 subjects who registered for the study, 124 who completed all prescreening, and 86 subjects who participated in centrifuge trials. Due to prior medical history, five subjects were disqualified, most commonly for psychiatric reasons or uncontrolled medical conditions. Of the subjects approved, four individuals experienced abnormal physiological responses to centrifuge profiles, including one back strain and three with anxiety reactions. The screening methods used were judged to be sufficient to identify individuals physically capable of tolerating simulated suborbital flight. Improved methods will be needed to identify susceptibility to anxiety reactions. While severe or uncontrolled disease was excluded, many subjects successfully participated in centrifuge trials despite medical histories of disease that are disqualifying under historical spaceflight screening regimes. Such screening techniques are applicable for use in future commercial spaceflight operations.

  5. SCALA: In situ calibration for integral field spectrographs

    Science.gov (United States)

    Lombardo, S.; Küsters, D.; Kowalski, M.; Aldering, G.; Antilogus, P.; Bailey, S.; Baltay, C.; Barbary, K.; Baugh, D.; Bongard, S.; Boone, K.; Buton, C.; Chen, J.; Chotard, N.; Copin, Y.; Dixon, S.; Fagrelius, P.; Feindt, U.; Fouchez, D.; Gangler, E.; Hayden, B.; Hillebrandt, W.; Hoffmann, A.; Kim, A. G.; Leget, P.-F.; McKay, L.; Nordin, J.; Pain, R.; Pécontal, E.; Pereira, R.; Perlmutter, S.; Rabinowitz, D.; Reif, K.; Rigault, M.; Rubin, D.; Runge, K.; Saunders, C.; Smadja, G.; Suzuki, N.; Taubenberger, S.; Tao, C.; Thomas, R. C.; Nearby Supernova Factory

    2017-11-01

    Aims: The scientific yield of current and future optical surveys is increasingly limited by systematic uncertainties in the flux calibration. This is the case for type Ia supernova (SN Ia) cosmology programs, where an improved calibration directly translates into improved cosmological constraints. Current methodology rests on models of stars. Here we aim to obtain flux calibration that is traceable to state-of-the-art detector-based calibration. Methods: We present the SNIFS Calibration Apparatus (SCALA), a color (relative) flux calibration system developed for the SuperNova integral field spectrograph (SNIFS), operating at the University of Hawaii 2.2 m (UH 88) telescope. Results: By comparing the color trend of the illumination generated by SCALA during two commissioning runs, and to previous laboratory measurements, we show that we can determine the light emitted by SCALA with a long-term repeatability better than 1%. We describe the calibration procedure necessary to control for system aging. We present measurements of the SNIFS throughput as estimated by SCALA observations. Conclusions: The SCALA calibration unit is now fully deployed at the UH 88 telescope, and with it color-calibration between 4000 Å and 9000 Å is stable at the percent level over a one-year baseline.

  6. Development of Camera Model and Geometric Calibration/validation of Xsat IRIS Imagery

    Science.gov (United States)

    Kwoh, L. K.; Huang, X.; Tan, W. J.

    2012-07-01

    XSAT, launched on 20 April 2011, is the first micro-satellite designed and built in Singapore. It orbits the Earth at altitude of 822 km in a sun synchronous orbit. The satellite carries a multispectral camera IRIS with three spectral bands - 0.52~0.60 mm for Green, 0.63~0.69 mm for Red and 0.76~0.89 mm for NIR at 12 m resolution. In the design of IRIS camera, the three bands were acquired by three lines of CCDs (NIR, Red and Green). These CCDs were physically separated in the focal plane and their first pixels not absolutely aligned. The micro-satellite platform was also not stable enough to allow for co-registration of the 3 bands with simple linear transformation. In the camera model developed, this platform stability was compensated with 3rd to 4th order polynomials for the satellite's roll, pitch and yaw attitude angles. With the camera model, the camera parameters such as the band to band separations, the alignment of the CCDs relative to each other, as well as the focal length of the camera can be validated or calibrated. The results of calibration with more than 20 images showed that the band to band along-track separation agreed well with the pre-flight values provided by the vendor (0.093° and 0.046° for the NIR vs red and for green vs red CCDs respectively). The cross-track alignments were 0.05 pixel and 5.9 pixel for the NIR vs red and green vs red CCDs respectively. The focal length was found to be shorter by about 0.8%. This was attributed to the lower operating temperature which XSAT is currently operating. With the calibrated parameters and the camera model, a geometric level 1 multispectral image with RPCs can be generated and if required, orthorectified imagery can also be produced.

  7. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  8. Gearbox Reliability Collaborative High-Speed Shaft Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Keller, J.; McNiff, B.

    2014-09-01

    Instrumentation has been added to the high-speed shaft, pinion, and tapered roller bearing pair of the Gearbox Reliability Collaborative gearbox to measure loads and temperatures. The new shaft bending moment and torque instrumentation was calibrated and the purpose of this document is to describe this calibration process and results, such that the raw shaft bending and torque signals can be converted to the proper engineering units and coordinate system reference for comparison to design loads and simulation model predictions.

  9. Initial inflight calibration for Hayabusa2 optical navigation camera (ONC) for science observations of asteroid Ryugu

    Science.gov (United States)

    Suzuki, H.; Yamada, M.; Kouyama, T.; Tatsumi, E.; Kameda, S.; Honda, R.; Sawada, H.; Ogawa, N.; Morota, T.; Honda, C.; Sakatani, N.; Hayakawa, M.; Yokota, Y.; Yamamoto, Y.; Sugita, S.

    2018-01-01

    Hayabusa2, the first sample return mission to a C-type asteroid was launched by the Japan Aerospace Exploration Agency (JAXA) on December 3, 2014 and will arrive at the asteroid in the middle of 2018 to collect samples from its surface, which may contain both hydrated minerals and organics. The optical navigation camera (ONC) system on board the Hayabusa2 consists of three individual framing CCD cameras, ONC-T for a telescopic nadir view, ONC-W1 for a wide-angle nadir view, and ONC-W2 for a wide-angle slant view will be used to observe the surface of Ryugu. The cameras will be used to measure the global asteroid shape, local morphologies, and visible spectroscopic properties. Thus, image data obtained by ONC will provide essential information to select landing (sampling) sites on the asteroid. This study reports the results of initial inflight calibration based on observations of Earth, Mars, Moon, and stars to verify and characterize the optical performance of the ONC, such as flat-field sensitivity, spectral sensitivity, point-spread function (PSF), distortion, and stray light of ONC-T, and distortion for ONC-W1 and W2. We found some potential problems that may influence our science observations. This includes changes in sensitivity of flat fields for all bands from those that were measured in the pre-flight calibration and existence of a stray light that arises under certain conditions of spacecraft attitude with respect to the sun. The countermeasures for these problems were evaluated by using data obtained during initial in-flight calibration. The results of our inflight calibration indicate that the error of spectroscopic measurements around 0.7 μm using 0.55, 0.70, and 0.86 μm bands of the ONC-T can be lower than 0.7% after these countermeasures and pixel binning. This result suggests that our ONC-T would be able to detect typical strength (∼3%) of the serpentine absorption band often found on CM chondrites and low albedo asteroids with ≥ 4

  10. Linear accelerator calibration monitor with a memory

    International Nuclear Information System (INIS)

    Dixon, R.L.; Ekstrand, K.E.

    1979-01-01

    A calibration monitor has been designed for measuring the constancy of linear accelerator or cobalt unit output between full calibrations. This monitor is battery-operated, light-weight, and slides into the shadow tray attachment on a linear accelerator or cobalt unit for easy setup. It provides a digital readout of the dose delivered, and a consistency check can be made in less than two minutes. The precision of the monitor, determined by cobalt-60 irradiations over a 2 1/2 month period, is +- 0.6% (standard deviation). The monitor also retains the dose reading in a CMOS digital counter indefinitely, hence it can be used in the same fashion as mailed thermoluminescent dosimeters (TLD) for calibration checks at remote facilities without the complicated readout procedures associated with TLD. The monitor can be mailed to a remote facility, positioned without ambiguity, and irradiated; and the reading can be verified on return to the originating center simply by pressing a switch. The monitor can easily be set up to carry out a ''blind'' check in which the reading obtained is not known to the remote facility

  11. Calibration and performance testing of the IAEA Aquila Active Well Coincidence Counter (Unit 1)

    International Nuclear Information System (INIS)

    Menlove, H.O..; Siebelist, R.; Wenz, T.R.

    1996-01-01

    An Active Well Coincidence Counter (AWCC) and a portable shift register (PSR-B) produced by Aquila Technologies Group, Inc., have been tested and cross-calibrated with existing AWCCs used by the International Atomic Energy Agency (IAEA). This report summarizes the results of these tests and the cross-calibration of the detector. In addition, updated tables summarizing the cross-calibration of existing AWCCs and AmLi sources are also included. Using the Aquila PSR-B with existing IAEA software requires secondary software also supplied by Aquila to set up the PSR-B with the appropriate measurement parameters

  12. PLEIADES ABSOLUTE CALIBRATION : INFLIGHT CALIBRATION SITES AND METHODOLOGY

    Directory of Open Access Journals (Sweden)

    S. Lachérade

    2012-07-01

    Full Text Available In-flight calibration of space sensors once in orbit is a decisive step to be able to fulfil the mission objectives. This article presents the methods of the in-flight absolute calibration processed during the commissioning phase. Four In-flight calibration methods are used: absolute calibration, cross-calibration with reference sensors such as PARASOL or MERIS, multi-temporal monitoring and inter-bands calibration. These algorithms are based on acquisitions over natural targets such as African deserts, Antarctic sites, La Crau (Automatic calibration station and Oceans (Calibration over molecular scattering or also new extra-terrestrial sites such as the Moon and selected stars. After an overview of the instrument and a description of the calibration sites, it is pointed out how each method is able to address one or several aspects of the calibration. We focus on how these methods complete each other in their operational use, and how they help building a coherent set of information that addresses all aspects of in-orbit calibration. Finally, we present the perspectives that the high level of agility of PLEIADES offers for the improvement of its calibration and a better characterization of the calibration sites.

  13. Calibration of the CREAM calorimeter with beam test data

    CERN Document Server

    Han, J H; Amare, Y

    The Cosmic Ray Energetics An d Mass (CREAM) calorimeter (CAL) is designed to measure cosmic-ray elemental energy spectra from 10 12 eV to 10 15 eV. It is comprised of 20 layers of tungsten interleaved with 20 layers of scintillating fiber ribbons. Before each flight, the CAL is exposed to an electron beam. For CREAM-IV through CREAM-VI, beams of 150 GeV electrons were used for the calibration, and 100 GeV was used for CREAM-VII. For calibration purpose, we compare electron beam data with simulation results to find calibration constants with the unit of MeV/ADC. In this paper, we present calibration results, including energy resolutions for electrons and uniformity of response. We also discuss CAL calibration using various beam test data compared with Monte Carlo (MC) simulation data.

  14. 3D-Calibration for IMU of the Strapdown Inertial Navigation Systems

    Directory of Open Access Journals (Sweden)

    Avrutov V.V.

    2017-01-01

    Full Text Available A new calibration method for Inertial Measurement Unit (IMU of Strapdown Iner-tial Navigation Systems was presented. IMU has been composed of accelerometers, gyroscopes and a circuit of signal processing. Normally, a rate transfer test and multi-position tests are us-ing for IMU calibration. The new calibration method is based on whole angle rotation or finite rotation. In fact it’s suggested to turn over IMU around three axes simultaneously. In order to solve the equation of calibration, it is necessary to provide an equality of a rank of basic matrix into degree of basic matrix. The results of simulated IMU data presented to demonstrate the performance of the new calibration method.

  15. Design, manufacture, and calibration of infrared radiometric blackbody sources

    International Nuclear Information System (INIS)

    Byrd, D.A.; Michaud, F.D.; Bender, S.C.

    1996-04-01

    A Radiometric Calibration Station (RCS) is being assembled at the Los Alamos National Laboratories (LANL) which will allow for calibration of sensors with detector arrays having spectral capability from about 0.4-15 μm. The configuration of the LANL RCS. Two blackbody sources have been designed to cover the spectral range from about 3-15 μm, operating at temperatures ranging from about 180-350 K within a vacuum environment. The sources are designed to present a uniform spectral radiance over a large area to the sensor unit under test. The thermal uniformity requirement of the blackbody cavities has been one of the key factors of the design, requiring less than 50 mK variation over the entire blackbody surface to attain effective emissivity values of about 0.999. Once the two units are built and verified to the level of about 100 mK at LANL, they will be sent to the National Institute of Standards and Technology (NIST), where at least a factor of two improvement will be calibrated into the blackbody control system. The physical size of these assemblies will require modifications of the existing NIST Low Background Infrared (LBIR) Facility. LANL has constructed a bolt-on addition to the LBIR facility that will allow calibration of our large aperture sources. Methodology for attaining the two blackbody sources at calibrated levels of performance equivalent to present state of the art will be explained in the following

  16. Uniformity calibration for ICT image

    International Nuclear Information System (INIS)

    Zeng Gang; Liu Li; Que Jiemin; Zhang Yingping; Yin Yin; Wang Yanfang; Yu Zhongqiang; Yan Yonglian

    2004-01-01

    The uniformity of ICT image is impaired by beam hardening and the inconsistency of detector units responses. The beam hardening and the nonlinearity of the detector's output have been analyzed. The correction factors are determined experimentally by the detector's responses with different absorption length. The artifacts in the CT image of a symmetrical aluminium cylinder have been eliminated after calibration. (author)

  17. 40 CFR 86.1319-90 - CVS calibration.

    Science.gov (United States)

    2010-07-01

    ....1319-90 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS (CONTINUED... “true” value) for the CVS calibration. (Note: In no case should an upstream screen or other restriction... Data Measurements Parameter Symbol Units Sensor-readout tolerances Barometric pressure (corrected) PB...

  18. Unit-specific calibration of Actigraph accelerometers in a mechanical setup - is it worth the effort? The effect on random output variation caused by technical inter-instrument variability in the laboratory and in the field

    DEFF Research Database (Denmark)

    Moeller, Niels C; Korsholm, Lars; Kristensen, Peter L

    2008-01-01

    BACKGROUND: Potentially, unit-specific in-vitro calibration of accelerometers could increase field data quality and study power. However, reduced inter-unit variability would only be important if random instrument variability contributes considerably to the total variation in field data. Therefor...

  19. Calibration

    International Nuclear Information System (INIS)

    Greacen, E.L.; Correll, R.L.; Cunningham, R.B.; Johns, G.G.; Nicolls, K.D.

    1981-01-01

    Procedures common to different methods of calibration of neutron moisture meters are outlined and laboratory and field calibration methods compared. Gross errors which arise from faulty calibration techniques are described. The count rate can be affected by the dry bulk density of the soil, the volumetric content of constitutional hydrogen and other chemical components of the soil and soil solution. Calibration is further complicated by the fact that the neutron meter responds more strongly to the soil properties close to the detector and source. The differences in slope of calibration curves for different soils can be as much as 40%

  20. An Improved Calibration Method for a Rotating 2D LIDAR System.

    Science.gov (United States)

    Zeng, Yadan; Yu, Heng; Dai, Houde; Song, Shuang; Lin, Mingqiang; Sun, Bo; Jiang, Wei; Meng, Max Q-H

    2018-02-07

    This paper presents an improved calibration method of a rotating two-dimensional light detection and ranging (R2D-LIDAR) system, which can obtain the 3D scanning map of the surroundings. The proposed R2D-LIDAR system, composed of a 2D LIDAR and a rotating unit, is pervasively used in the field of robotics owing to its low cost and dense scanning data. Nevertheless, the R2D-LIDAR system must be calibrated before building the geometric model because there are assembled deviation and abrasion between the 2D LIDAR and the rotating unit. Hence, the calibration procedures should contain both the adjustment between the two devices and the bias of 2D LIDAR itself. The main purpose of this work is to resolve the 2D LIDAR bias issue with a flat plane based on the Levenberg-Marquardt (LM) algorithm. Experimental results for the calibration of the R2D-LIDAR system prove the reliability of this strategy to accurately estimate sensor offsets with the error range from -15 mm to 15 mm for the performance of capturing scans.

  1. Calibration of dosemeters and survey instruments for photon

    International Nuclear Information System (INIS)

    Alamares, A.L.; Caseria, E.S.

    1995-01-01

    The Philippine Nuclear Research Institute (PNRI), the country's regulatory agency, through its Radiation Protection Unit of the Nuclear Regulations Licensing and Safeguard Division provides the calibration services of nuclear instruments used by various institutions. The periodic calibration and standardization of radiation and protection survey instruments are done to insure correct and valid radiation readings. It is also a regulatory requirement for radiation workers to use only operable and precalibrated survey instruments for their work with radioactive materials. The PNRI maintains and operates the Secondary Standard Dosimetry Laboratory (SSDL). It provides the periodic calibration and standardization of radiation monitoring and protection survey instruments, and also the radiological hazard and performance evaluation surveys of radiotherapy facilities and others. For the calibration of survey instruments, the standard sources of Cs-137 and Co-60 are used. The calibration procedures for survey meters and pocket dosimeters are explained. The clients served are shown. The personnel dosimetry using film badges, the application of ICRU operational quantities, and the problems encountered and recommendations are reported. (K.I.)

  2. Traceable calibration of photovoltaic reference cells using natural sunlight

    Science.gov (United States)

    Müllejans, H.; Zaaiman, W.; Pavanello, D.; Dunlop, E. D.

    2018-02-01

    At the European Solar Test Installation (ESTI) photovoltaic (PV) reference cells are calibrated traceably to SI units via the World Radiometric Reference (WRR) using natural sunlight. The Direct Sunlight Method (DSM) is described in detail and the latest measurement results and an updated uncertainty budget are reported. These PV reference cells then provide a practical means for measuring the irradiance of natural or simulated sunlight during the calibration of other PV devices.

  3. Calibration procedures for mammography dosemeters in Poland

    International Nuclear Information System (INIS)

    Gwiazdowska, B.; Ulkowski, P.; Tolwinski, J.; Bulski, W.

    2002-01-01

    Breast cancer is the most frequent tumour in women and the effectiveness of the treatment depends dramatically on the early detection of the disease. That is the reason why in Poland the mammography control examinations are strongly supported by the Centre of Oncology. In Poland there are over 400 mammography units which account for about 300,000 examinations per year. An investigation performed by the Medical Physics Department of the Centre of Oncology in Warsaw at about 100 mammography facilities proved that in most cases the doses absorbed by the patients could be reduced without decrease of image quality. This is one of the reasons why the Polish Secondary Standard Dosimetry Laboratory (SSDL) dealing mainly with calibration of radiotherapy dosemeters is extending its activities and therefore new facilities and equipment adapted for calibration of mammographic dosemeters have been installed. The mammography dosimetry calibration equipment is permanently installed in the same laboratory room where the radiotherapy dosemeters are calibrated. A base of a mammography unit no longer in clinical use, together with its movable system has been adapted to handle ionization chamber holders. An X-ray tube with a 50 kV high frequency generator was also installed. The tube, a Varian type OEG-50-2, (designed for laboratory applications) with molybdenum anode of an anode angle 23,7 deg. and with a large focus, effective size approximately 5 mm 2 , has an inherent filtration of 0,25 mm beryllium. It is installed in a housing with 2mm lead shielding; a cone shaped beam is formed by a system of three collimators

  4. Development of nano-roughness calibration standards

    International Nuclear Information System (INIS)

    Baršić, Gorana; Mahović, Sanjin; Zorc, Hrvoje

    2012-01-01

    At the Laboratory for Precise Measurements of Length, currently the Croatian National Laboratory for Length, unique nano-roughness calibration standards were developed, which have been physically implemented in cooperation with the company MikroMasch Trading OU and the Ruđer Bošković Institute. In this paper, a new design for a calibration standard with two measuring surfaces is presented. One of the surfaces is for the reproduction of roughness parameters, while the other is for the traceability of length units below 50 nm. The nominal values of the groove depths on these measuring surfaces are the same. Thus, a link between the measuring surfaces has been ensured, which makes these standards unique. Furthermore, the calibration standards available on the market are generally designed specifically for individual groups of measuring instrumentation, such as interferometric microscopes, stylus instruments, scanning electron microscopes (SEM) or scanning probe microscopes. In this paper, a new design for nano-roughness standards has been proposed for use in the calibration of optical instruments, as well as for stylus instruments, SEM, atomic force microscopes and scanning tunneling microscopes. Therefore, the development of these new nano-roughness calibration standards greatly contributes to the reproducibility of the results of groove depth measurement as well as the 2D and 3D roughness parameters obtained by various measuring methods. (paper)

  5. Measurement of TLD Albedo response on various calibration phantoms

    International Nuclear Information System (INIS)

    Momose, T.; Tsujimura, N.; Shinohara, K.; Ishiguro, H.; Nakamura, T.

    1996-01-01

    The International Commission on Radiation Units and Measurements (ICRU) has recommended that individual dosemeter should be calibrated on a suitable phantom and has pointed out that the calibration factor of a neutron dosemeter is strongly influenced by the the exact size and shape of the body and the phantom to which the dosemeter is attached. As the principle of an albedo type thermoluminescent personal dosemeter (albedo TLD) is essentially based on a detection of scattered and moderated neutron from a human body, the sensitivity of albedo TLD is strongly influenced by the incident neutron energy and the calibration phantom. (1) Therefore for albedo type thermoluminescent personal dosemeter (albedo TLD), the information of neutron albedo response on the calibration phantom is important for appropriate dose estimation. In order to investigate the effect of phantom type on the reading of the albedo TLD, measurement of the TLD energy response and angular response on some typical calibration phantoms was performed using dynamitron accelerator and 252 Cf neutron source. (author)

  6. Gearbox Reliability Collaborative Gearbox 3 Planet Bearing Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Keller, Jonathan [National Renewable Energy Lab. (NREL), Golden, CO (United States)

    2017-03-24

    The Gearbox Reliability Collaborative gearbox was redesigned to improve its load-sharing characteristics and predicted fatigue life. The most important aspect of the redesign was to replace the cylindrical roller bearings with preloaded tapered roller bearings in the planetary section. Similar to previous work, the strain gages installed on the planet tapered roller bearings were calibrated in a load frame. This report describes the calibration tests and provides the factors necessary to convert the measured units from dynamometer testing to bearing loads, suitable for comparison to engineering models.

  7. A Monte Carlo modeling alternative for the API Gamma Ray Calibration Facility

    International Nuclear Information System (INIS)

    Galford, J.E.

    2017-01-01

    The gamma ray pit at the API Calibration Facility, located on the University of Houston campus, defines the API unit for natural gamma ray logs used throughout the petroleum logging industry. Future use of the facility is uncertain. An alternative method is proposed to preserve the gamma ray API unit definition as an industry standard by using Monte Carlo modeling to obtain accurate counting rate-to-API unit conversion factors for gross-counting and spectral gamma ray tool designs. - Highlights: • A Monte Carlo alternative is proposed to replace empirical calibration procedures. • The proposed Monte Carlo alternative preserves the original API unit definition. • MCNP source and materials descriptions are provided for the API gamma ray pit. • Simulated results are presented for several wireline logging tool designs. • The proposed method can be adapted for use with logging-while-drilling tools.

  8. IMU-based online kinematic calibration of robot manipulator.

    Science.gov (United States)

    Du, Guanglong; Zhang, Ping

    2013-01-01

    Robot calibration is a useful diagnostic method for improving the positioning accuracy in robot production and maintenance. An online robot self-calibration method based on inertial measurement unit (IMU) is presented in this paper. The method requires that the IMU is rigidly attached to the robot manipulator, which makes it possible to obtain the orientation of the manipulator with the orientation of the IMU in real time. This paper proposed an efficient approach which incorporates Factored Quaternion Algorithm (FQA) and Kalman Filter (KF) to estimate the orientation of the IMU. Then, an Extended Kalman Filter (EKF) is used to estimate kinematic parameter errors. Using this proposed orientation estimation method will result in improved reliability and accuracy in determining the orientation of the manipulator. Compared with the existing vision-based self-calibration methods, the great advantage of this method is that it does not need the complex steps, such as camera calibration, images capture, and corner detection, which make the robot calibration procedure more autonomous in a dynamic manufacturing environment. Experimental studies on a GOOGOL GRB3016 robot show that this method has better accuracy, convenience, and effectiveness than vision-based methods.

  9. IMU-Based Online Kinematic Calibration of Robot Manipulator

    Directory of Open Access Journals (Sweden)

    Guanglong Du

    2013-01-01

    Full Text Available Robot calibration is a useful diagnostic method for improving the positioning accuracy in robot production and maintenance. An online robot self-calibration method based on inertial measurement unit (IMU is presented in this paper. The method requires that the IMU is rigidly attached to the robot manipulator, which makes it possible to obtain the orientation of the manipulator with the orientation of the IMU in real time. This paper proposed an efficient approach which incorporates Factored Quaternion Algorithm (FQA and Kalman Filter (KF to estimate the orientation of the IMU. Then, an Extended Kalman Filter (EKF is used to estimate kinematic parameter errors. Using this proposed orientation estimation method will result in improved reliability and accuracy in determining the orientation of the manipulator. Compared with the existing vision-based self-calibration methods, the great advantage of this method is that it does not need the complex steps, such as camera calibration, images capture, and corner detection, which make the robot calibration procedure more autonomous in a dynamic manufacturing environment. Experimental studies on a GOOGOL GRB3016 robot show that this method has better accuracy, convenience, and effectiveness than vision-based methods.

  10. A Start-up Calibration Method for Generic Current-Steering D/A Converters with Optimal Area Solution

    NARCIS (Netherlands)

    Radulov, G.I.; Quinn, P.J.; Hegt, J.A.; Roermund, van A.H.M.

    2005-01-01

    This paper presents a new start-up calibration method for current-steering D/A converters, based on a 1-bit ADC. The paper proposes a new current cell that allows calibration of non-identical current sources by way of a shared calibration apparatus. The current cell uses parallel selfcalibrated unit

  11. Antenna unit and radio base station therewith

    Science.gov (United States)

    Kuwahara, Mikio; Doi, Nobukazu; Suzuki, Toshiro; Ishida, Yuji; Inoue, Takashi; Niida, Sumaru

    2007-04-10

    Phase and amplitude deviations, which are generated, for example, by cables connecting an array antenna of a CDMA base station and the base station, are calibrated in the baseband. The base station comprises: an antenna apparatus 1; couplers 2; an RF unit 3 that converts a receive signal to a baseband signal, converts a transmit signal to a radio frequency, and performs power control; an A/D converter 4 for converting a receive signal to a digital signal; a receive beam form unit 6 that multiplies the receive signal by semi-fixed weight; a despreader 7 for this signal input; a time-space demodulator 8 for demodulating user data; a despreader 9 for probe signal; a space modulator 14 for user data; a spreader 13 for user signal; a channel combiner 12; a Tx calibrater 11 for controlling calibration of a signal; a D/A converter 10; a unit 16 for calculation of correlation matrix for generating a probe signal used for controlling an Rx calibration system and a TX calibration system; a spreader 17 for probe signal; a power control unit 18; a D/A converter 19; an RF unit 20 for probe signal; an A/D converter 21 for signal from the couplers 2; and a despreader 22.

  12. An Improved Calibration Method for a Rotating 2D LIDAR System

    Directory of Open Access Journals (Sweden)

    Yadan Zeng

    2018-02-01

    Full Text Available This paper presents an improved calibration method of a rotating two-dimensional light detection and ranging (R2D-LIDAR system, which can obtain the 3D scanning map of the surroundings. The proposed R2D-LIDAR system, composed of a 2D LIDAR and a rotating unit, is pervasively used in the field of robotics owing to its low cost and dense scanning data. Nevertheless, the R2D-LIDAR system must be calibrated before building the geometric model because there are assembled deviation and abrasion between the 2D LIDAR and the rotating unit. Hence, the calibration procedures should contain both the adjustment between the two devices and the bias of 2D LIDAR itself. The main purpose of this work is to resolve the 2D LIDAR bias issue with a flat plane based on the Levenberg–Marquardt (LM algorithm. Experimental results for the calibration of the R2D-LIDAR system prove the reliability of this strategy to accurately estimate sensor offsets with the error range from −15 mm to 15 mm for the performance of capturing scans.

  13. A SAR-ADC using unit bridge capacitor and with calibration for the front-end electronics of PET imaging

    Energy Technology Data Exchange (ETDEWEB)

    Liu, Wei [School of Computer Science and Engineering, Northwestern Polytechnical University, Xi' an 710072, Shaanxi (China); Wei, Tingcun, E-mail: weitc@nwpu.edu.cn [School of Computer Science and Engineering, Northwestern Polytechnical University, Xi' an 710072, Shaanxi (China); Li, Bo; Yang, Lifeng; Xue, Feifei [School of Computer Science and Engineering, Northwestern Polytechnical University, Xi' an 710072, Shaanxi (China); Hu, Yongcai [Institut Pluridisciplinaire Hubert CURIEN, Strasbourg (France)

    2016-05-11

    This paper presents a 12-bit 1 MS/s successive approximation register-analog to digital converter (SAR-ADC) for the 32-channel front-end electronics of CZT-based PET imaging system. To reduce the capacitance mismatch, instead of the fractional capacitor, the unit capacitor is used as the bridge capacitor in the split-capacitor digital to analog converter (DAC) circuit. In addition, in order to eliminate the periodical DNL errors of −1 LSB which often exists in the SAR-ADC using the charge-redistributed DAC, a calibration algorithm is proposed and verified by the experiments. The proposed 12-bit 1 MS/s SAR-ADC is designed and implemented using a 0.35 μm CMOS technology, it occupies only an active area of 986×956 μm{sup 2}. The measurement results show that, at the power supply of 3.3/5.0 V and the sampling rate of 1 MS/s, the ADC with calibration has a signal-to-noise-and-distortion ratio (SINAD) of 67.98 dB, the power dissipation of 5 mW, and a figure of merit (FOM) of 2.44 pJ/conv.-step. This ADC is with the features of high accuracy, low power and small layout area, it is especially suitable to the one-chip integration of the front-end readout electronics.

  14. A new systematic calibration method of ring laser gyroscope inertial navigation system

    Science.gov (United States)

    Wei, Guo; Gao, Chunfeng; Wang, Qi; Wang, Qun; Xiong, Zhenyu; Long, Xingwu

    2016-10-01

    Inertial navigation system has been the core component of both military and civil navigation systems. Before the INS is put into application, it is supposed to be calibrated in the laboratory in order to compensate repeatability error caused by manufacturing. Discrete calibration method cannot fulfill requirements of high-accurate calibration of the mechanically dithered ring laser gyroscope navigation system with shock absorbers. This paper has analyzed theories of error inspiration and separation in detail and presented a new systematic calibration method for ring laser gyroscope inertial navigation system. Error models and equations of calibrated Inertial Measurement Unit are given. Then proper rotation arrangement orders are depicted in order to establish the linear relationships between the change of velocity errors and calibrated parameter errors. Experiments have been set up to compare the systematic errors calculated by filtering calibration result with those obtained by discrete calibration result. The largest position error and velocity error of filtering calibration result are only 0.18 miles and 0.26m/s compared with 2 miles and 1.46m/s of discrete calibration result. These results have validated the new systematic calibration method and proved its importance for optimal design and accuracy improvement of calibration of mechanically dithered ring laser gyroscope inertial navigation system.

  15. Skew redundant MEMS IMU calibration using a Kalman filter

    International Nuclear Information System (INIS)

    Jafari, M; Sahebjameyan, M; Moshiri, B; Najafabadi, T A

    2015-01-01

    In this paper, a novel calibration procedure for skew redundant inertial measurement units (SRIMUs) based on micro-electro mechanical systems (MEMS) is proposed. A general model of the SRIMU measurements is derived which contains the effects of bias, scale factor error and misalignments. For more accuracy, the effect of lever arms of the accelerometers to the center of the table are modeled and compensated in the calibration procedure. Two separate Kalman filters (KFs) are proposed to perform the estimation of error parameters for gyroscopes and accelerometers. The predictive error minimization (PEM) stochastic modeling method is used to simultaneously model the effect of bias instability and random walk noise on the calibration Kalman filters to diminish the biased estimations. The proposed procedure is simulated numerically and has expected experimental results. The calibration maneuvers are applied using a two-axis angle turntable in a way that the persistency of excitation (PE) condition for parameter estimation is met. For this purpose, a trapezoidal calibration profile is utilized to excite different deterministic error parameters of the accelerometers and a pulse profile is used for the gyroscopes. Furthermore, to evaluate the performance of the proposed KF calibration method, a conventional least squares (LS) calibration procedure is derived for the SRIMUs and the simulation and experimental results compare the functionality of the two proposed methods with each other. (paper)

  16. New approach for calibration and interpretation of IRAD GAGE vibrating-wire stressmeters

    International Nuclear Information System (INIS)

    Mao, N.

    1986-05-01

    IRAD GAGE vibrating-wire stressmeters were installed in the Spent Fuel Facility at the Nevada Test Site to measure the change in in-situ stress during the Spent Fuel Test-Climax (SFT-C). This paper discusses the results of removing a cylindrical section of rock and gages as a unit through overcoring, and the subsequent post-test calibration of the stressmeters in the laboratory. The estimated in-situ stresses based on post test calibration data are quite consistent with those directly measured in nearby holes. The magnitude of stress change calculated from pre-test calibration data is generally much smaller than that estimated from post test calibration data. 11 refs., 5 figs., 2 tabs

  17. Total-Count Calibration Blocks for use in uranium Exploration

    DEFF Research Database (Denmark)

    Løvborg, Leif

    Transportable calibration blocks for field scintillometers and borehole probes were manufactured from concrete and installed at calibration sites in Denmark and Greece. The concrete mixes were prepared from aggregates of quartz sand and crushed uranium-thorium ore. Hater-reducing agents and silica...... dust added to the cement paste produced concretes of acceptable porosity and pore structure. The content of ore was adjusted to provide block grades of approximately 2, 140, and 540 units of radioelement concentration (Ur). Thorium was estimated to contribute 0.39 ± 0.02 Ur per ppm Th. The adopted...

  18. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    International Nuclear Information System (INIS)

    Cornic, Philippe; Le Besnerais, Guy; Champagnat, Frédéric; Illoul, Cédric; Cheminet, Adam; Le Sant, Yves; Leclaire, Benjamin

    2016-01-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data. (paper)

  19. Calibration of robot tool centre point using camera-based system

    Directory of Open Access Journals (Sweden)

    Gordić Zaviša

    2016-01-01

    Full Text Available Robot Tool Centre Point (TCP calibration problem is of great importance for a number of industrial applications, and it is well known both in theory and in practice. Although various techniques have been proposed for solving this problem, they mostly require tool jogging or long processing time, both of which affect process performance by extending cycle time. This paper presents an innovative way of TCP calibration using a set of two cameras. The robot tool is placed in an area where images in two orthogonal planes are acquired using cameras. Using robust pattern recognition, even deformed tool can be identified on images, and information about its current position and orientation forwarded to control unit for calibration. Compared to other techniques, test results show significant reduction in procedure complexity and calibration time. These improvements enable more frequent TCP checking and recalibration during production, thus improving the product quality.

  20. Design and Implementation of High Precision Temperature Measurement Unit

    Science.gov (United States)

    Zeng, Xianzhen; Yu, Weiyu; Zhang, Zhijian; Liu, Hancheng

    2018-03-01

    Large-scale neutrino detector requires calibration of photomultiplier tubes (PMT) and electronic system in the detector, performed by plotting the calibration source with a group of designated coordinates in the acrylic sphere. Where the calibration source positioning is based on the principle of ultrasonic ranging, the transmission speed of ultrasonic in liquid scintillator of acrylic sphere is related to temperature. This paper presents a temperature measurement unit based on STM32L031 and single-line bus digital temperature sensor TSic506. The measurement data of the temperature measurement unit can help the ultrasonic ranging to be more accurate. The test results show that the temperature measurement error is within ±0.1°C, which satisfies the requirement of calibration source positioning. Take energy-saving measures, with 3.7V/50mAH lithium battery-powered, the temperature measurement unit can work continuously more than 24 hours.

  1. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Joseph Rovani; Mark Sanderson

    2008-02-29

    Mercury continuous emissions monitoring systems (CEMS) are being implemented in over 800 coal-fired power plant stacks. The power industry desires to conduct at least a full year of monitoring before the formal monitoring and reporting requirement begins on January 1, 2009. It is important for the industry to have available reliable, turnkey equipment from CEM vendors. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The generators are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 requires that calibration be performed with NIST-traceable standards (Federal Register 2007). Traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued an interim traceability protocol for elemental mercury generators (EPA 2007). The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The document is divided into two separate sections. The first deals with the qualification of generators by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the generator models that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma/mass spectrometry performed by NIST in Gaithersburg, MD. The

  2. Radiation Measurements Laboratory (RML) calibration and assessment of the ATR SPING-3 stack effluent monitor

    International Nuclear Information System (INIS)

    Koeppen, L.D.; Rogers, J.W.; Simpson, O.D.

    1983-12-01

    An evaluation, calibration and assessment of the Eberline SPING-3 ATR stack effluent monitor was conducted. This unit which monitors particulate, iodine and noble gas effluents was producing abnormal results following the initial installation and operational testing. The purposes of this work were to find the causes of the abnormal results and correct them if possible; check the calibrations and adjust them if necessary; and to provide a better in-depth understanding of what the unit is monitoring and how well it performs under this application. Results have shown that there were some problems associated with the unit as initially installed and tested. These problems have been identified and suggested alternatives shown, the monitor was found to be applicable to some extent under the current conditions. The calibrations have been checked and adjustments made. More operation testing and evaluation is needed to assess how well this works under a variety of ATR operating conditions. 2 references, 10 figures, 3 tables

  3. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration...

  4. In-flight calibration of the ISGRI camera

    International Nuclear Information System (INIS)

    Terrier, R.; Lebrun, F.; Belanger, G.; Blondel, C.; David, P.; Goldoni, P.; Goldwurm, A.; Gros, A.; Laurent, P.; Malaguti, G.; Sauvageon, A.; Bazzano, A.; Ubertini, P.; Segreto, A.; Malaguti, G.; Bird, A.J.

    2003-01-01

    ISGRI, the IBIS low energy camera (15 keV-1 MeV) on board INTEGRAL, is the first large CdTe gamma-ray imager in orbit. We present here an overview of the ISGRI in-flight calibrations performed during the first months after launch. We discuss the stability of the camera as well as the CdTe pixels response under cosmic radiation. The energy calibrations were done using lead and tungsten fluorescence lines and the 22 Na calibration unit. Thermal effects and charge correction algorithm are discussed, and the resulting energy resolution is presented. The ISGRI background spatial and spectral non-uniformity is also described, and some image correction results are presented. ISGRI, despite a few unexpected features like zero rise time events, performs well with only 4,5% noisy or disabled pixels. Thermal effects are at the origin of the largest difference between ground and in-flight data. Correcting for these effects yields good spectral performances close to the expectations with 8.4% at 59.3 keV and 4.9% at 511 keV. The resolution in the high energy band is broader than before launch because of residual rise time gains uncertainties. Handling of these errors requires a larger amount of calibration data than what is available today

  5. A Synthesis of Star Calibration Techniques for Ground-Based Narrowband Electron-Multiplying Charge-Coupled Device Imagers Used in Auroral Photometry

    Science.gov (United States)

    Grubbs, Guy II; Michell, Robert; Samara, Marilia; Hampton, Don; Jahn, Jorg-Micha

    2016-01-01

    A technique is presented for the periodic and systematic calibration of ground-based optical imagers. It is important to have a common system of units (Rayleighs or photon flux) for cross comparison as well as self-comparison over time. With the advancement in technology, the sensitivity of these imagers has improved so that stars can be used for more precise calibration. Background subtraction, flat fielding, star mapping, and other common techniques are combined in deriving a calibration technique appropriate for a variety of ground-based imager installations. Spectral (4278, 5577, and 8446 A ) ground-based imager data with multiple fields of view (19, 47, and 180 deg) are processed and calibrated using the techniques developed. The calibration techniques applied result in intensity measurements in agreement between different imagers using identical spectral filtering, and the intensity at each wavelength observed is within the expected range of auroral measurements. The application of these star calibration techniques, which convert raw imager counts into units of photon flux, makes it possible to do quantitative photometry. The computed photon fluxes, in units of Rayleighs, can be used for the absolute photometry between instruments or as input parameters for auroral electron transport models.

  6. Calibrated infrared ground/air radiometric spectrometer

    Science.gov (United States)

    Silk, J. K.; Schildkraut, Elliot Robert; Bauldree, Russell S.; Goodrich, Shawn M.

    1996-06-01

    The calibrated infrared ground/air radiometric spectrometer (CIGARS) is a new high performance, multi-purpose, multi- platform Fourier transform spectrometer (FPS) sensor. It covers the waveband from 0.2 to 12 micrometer, has spectral resolution as fine as 0.3 cm-1, and records over 100 spectra per second. Two CIGARS units are being used for observations of target signatures in the air or on the ground from fixed or moving platforms, including high performance jet aircraft. In this paper we describe the characteristics and capabilities of the CIGARS sensor, which uses four interchangeable detector modules (Si, InGaAs, InSb, and HgCdTe) and two optics modules, with internal calibration. The data recording electronics support observations of transient events, even without precise information on the timing of the event. We present test and calibration data on the sensitivity, spectral resolution, stability, and spectral rate of CIGARS, and examples of in- flight observations of real targets. We also discuss plans for adapting CIGARS for imaging spectroscopy observations, with simultaneous spectral and spatial data, by replacing the existing detectors with a focal plane array (FPA).

  7. Development of the laser alignment system with PSD used for shaft calibration

    Science.gov (United States)

    Jiao, Guohua; Li, Yulin; Hu, Baowen

    2006-02-01

    Shaft calibration is an important technique during installation and maintenance of a rotating machine. It requires unique and high-precision measurement instruments with calculation capability, and relies on experience on heavy, high-speed, or high-temperature machines. A high-precision laser alignment system has been designed using PSD (Position Sensing Detector) to change traditional manual way of shaft calibration and to make the measurement easier and more accurate. The system is comprised of two small measuring units (Laser transmitter and detector) and a hand operated control unit or a PC. Such a laser alignment system has been used in some actual shaft alignment with offset resolution 1.5μm and angular resolution 0.1°.

  8. Development of a quality assurance program for ionizing radiation secondary calibration laboratories

    Energy Technology Data Exchange (ETDEWEB)

    Heaton, H.T. II; Taylor, A.R. Jr. [Center for Devices and Radiological Health, Rockville, MD (United States)

    1993-12-31

    For calibration laboratories, routine calibrations of instruments meeting stated accuracy goals are important. One method of achieving the accuracy goals is to establish and follow a quality assurance program designed to monitor all aspects of the calibration program and to provide the appropriate feedback mechanism if adjustments are needed. In the United States there are a number of organizations with laboratory accreditation programs. All existing accreditation programs require that the laboratory implement a quality assurance program with essentially the same elements in all of these programs. Collectively, these elements have been designated as a Measurement Quality Assurance (MQA) program. This paper will briefly discuss the interrelationship of the elements of an MQA program. Using the Center for Devices and Radiological Health (CDRH) X-ray Calibration Laboratory (XCL) as an example, it will focus on setting up a quality control program for the equipment in a Secondary Calibration Laboratory.

  9. Development of a quality assurance program for ionizing radiation secondary calibration laboratories

    International Nuclear Information System (INIS)

    Heaton, H.T. II; Taylor, A.R. Jr.

    1993-01-01

    For calibration laboratories, routine calibrations of instruments meeting stated accuracy goals are important. One method of achieving the accuracy goals is to establish and follow a quality assurance program designed to monitor all aspects of the calibration program and to provide the appropriate feedback mechanism if adjustments are needed. In the United States there are a number of organizations with laboratory accreditation programs. All existing accreditation programs require that the laboratory implement a quality assurance program with essentially the same elements in all of these programs. Collectively, these elements have been designated as a Measurement Quality Assurance (MQA) program. This paper will briefly discuss the interrelationship of the elements of an MQA program. Using the Center for Devices and Radiological Health (CDRH) X-ray Calibration Laboratory (XCL) as an example, it will focus on setting up a quality control program for the equipment in a Secondary Calibration Laboratory

  10. Exposure-rate calibration using large-area calibration pads

    International Nuclear Information System (INIS)

    Novak, E.F.

    1988-09-01

    The US Department of Energy (DOE) Office of Remedial Action and Waste Technology established the Technical Measurements Center (TMC) at the DOE Grand Junction Projects Office (GJPO) in Grand Junction, Colorado, to standardize, calibrate, and compare measurements made in support of DOE remedial action programs. A set of large-area, radioelement-enriched concrete pads was constructed by the DOE in 1978 at the Walker Field Airport in Grand Junction for use as calibration standards for airborne gamma-ray spectrometer systems. The use of these pads was investigated by the TMC as potential calibration standards for portable scintillometers employed in measuring gamma-ray exposure rates at Uranium Mill Tailings Remedial Action (UMTRA) project sites. Data acquired on the pads using a pressurized ionization chamber (PIC) and three scintillometers are presented as an illustration of an instrumental calibration. Conclusions and recommended calibration procedures are discussed, based on the results of these data

  11. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  12. Generator Dynamic Model Validation and Parameter Calibration Using Phasor Measurements at the Point of Connection

    Energy Technology Data Exchange (ETDEWEB)

    Huang, Zhenyu; Du, Pengwei; Kosterev, Dmitry; Yang, Steve

    2013-05-01

    Disturbance data recorded by phasor measurement units (PMU) offers opportunities to improve the integrity of dynamic models. However, manually tuning parameters through play-back events demands significant efforts and engineering experiences. In this paper, a calibration method using the extended Kalman filter (EKF) technique is proposed. The formulation of EKF with parameter calibration is discussed. Case studies are presented to demonstrate its validity. The proposed calibration method is cost-effective, complementary to traditional equipment testing for improving dynamic model quality.

  13. South African initiative for pre-flight radiometric calibration of satellite imagers

    CSIR Research Space (South Africa)

    Griffith, D

    2009-07-01

    Full Text Available -Thomson prism. The resulting beam should be depolarized either with an optically thick diffuser (e.g. flashed opal) or with a pseudo-depolarizer if the loss of light at such a diffuser is unacceptable. 5. SUMMARY The above described South African...

  14. Principles and applications of measurement and uncertainty analysis in research and calibration

    Energy Technology Data Exchange (ETDEWEB)

    Wells, C.V.

    1992-11-01

    Interest in Measurement Uncertainty Analysis has grown in the past several years as it has spread to new fields of application, and research and development of uncertainty methodologies have continued. This paper discusses the subject from the perspectives of both research and calibration environments. It presents a history of the development and an overview of the principles of uncertainty analysis embodied in the United States National Standard, ANSI/ASME PTC 19.1-1985, Measurement Uncertainty. Examples are presented in which uncertainty analysis was utilized or is needed to gain further knowledge of a particular measurement process and to characterize final results. Measurement uncertainty analysis provides a quantitative estimate of the interval about a measured value or an experiment result within which the true value of that quantity is expected to lie. Years ago, Harry Ku of the United States National Bureau of Standards stated that ``The informational content of the statement of uncertainty determines, to a large extent, the worth of the calibrated value.`` Today, that statement is just as true about calibration or research results as it was in 1968. Why is that true? What kind of information should we include in a statement of uncertainty accompanying a calibrated value? How and where do we get the information to include in an uncertainty statement? How should we interpret and use measurement uncertainty information? This discussion will provide answers to these and other questions about uncertainty in research and in calibration. The methodology to be described has been developed by national and international groups over the past nearly thirty years, and individuals were publishing information even earlier. Yet the work is largely unknown in many science and engineering arenas. I will illustrate various aspects of uncertainty analysis with some examples drawn from the radiometry measurement and calibration discipline from research activities.

  15. Principles and applications of measurement and uncertainty analysis in research and calibration

    Energy Technology Data Exchange (ETDEWEB)

    Wells, C.V.

    1992-11-01

    Interest in Measurement Uncertainty Analysis has grown in the past several years as it has spread to new fields of application, and research and development of uncertainty methodologies have continued. This paper discusses the subject from the perspectives of both research and calibration environments. It presents a history of the development and an overview of the principles of uncertainty analysis embodied in the United States National Standard, ANSI/ASME PTC 19.1-1985, Measurement Uncertainty. Examples are presented in which uncertainty analysis was utilized or is needed to gain further knowledge of a particular measurement process and to characterize final results. Measurement uncertainty analysis provides a quantitative estimate of the interval about a measured value or an experiment result within which the true value of that quantity is expected to lie. Years ago, Harry Ku of the United States National Bureau of Standards stated that The informational content of the statement of uncertainty determines, to a large extent, the worth of the calibrated value.'' Today, that statement is just as true about calibration or research results as it was in 1968. Why is that true What kind of information should we include in a statement of uncertainty accompanying a calibrated value How and where do we get the information to include in an uncertainty statement How should we interpret and use measurement uncertainty information This discussion will provide answers to these and other questions about uncertainty in research and in calibration. The methodology to be described has been developed by national and international groups over the past nearly thirty years, and individuals were publishing information even earlier. Yet the work is largely unknown in many science and engineering arenas. I will illustrate various aspects of uncertainty analysis with some examples drawn from the radiometry measurement and calibration discipline from research activities.

  16. IAEA workshop/seminar on calibration procedures in dosimetry, Quito, 6-24 October 1986

    International Nuclear Information System (INIS)

    1987-01-01

    The International Atomic Energy Agency in co-operation with the Ecuadorian Atomic Energy Commission organized a workshop and seminar on calibration procedures in dosimetry at the SSDL Quito, 6 to 24 October 1986. All calibration laboratories in the Latin American region were invited to participate. The purpose of the workshop were calibration exercises with therapy-level and protection-level secondary standards at various calibration qualities, discussions on progress made in the different SSDLs in the region and delivering lectures on pertinent subjects. A total of 15 Secondary Standards (10 therapy-level and 5 protection-level) were brought along by the participants and 35 calibration comparisons were performed with those having a valid calibration factor. Thirty-three determinations of calibration factors were performed for secondary standards not having had a calibration before. Twelve different calibration qualities were available (Cobalt-60 and X-rays) and Agency's Secondary Standards traceable to BIPM were the reference standards. The participants were divided into two working groups, each one week and each group into two sub-groups. Both irradiation bunkers were used simultaneously. The one houses the teletherapy Cobalt-60 unit and the protection-level Cobalt-60 irradiator, the other one the constant potential X-ray machine with maximum generating potential of 320 KV and suitable for both therapy-level as well as protection-level calibrations. Due to the heavy workload and limited time available some nightshifts were required to accomplish the requested calibration comparisons

  17. Automated calibration system for a high-precision measurement of neutrino mixing angle θ13 with the Daya Bay antineutrino detectors

    International Nuclear Information System (INIS)

    Liu, J.; Cai, B.; Carr, R.; Dwyer, D.A.; Gu, W.Q.; Li, G.S.; Qian, X.; McKeown, R.D.; Tsang, R.H.M.; Wang, W.; Wu, F.F.; Zhang, C.

    2014-01-01

    We describe the automated calibration system for the antineutrino detectors in the Daya Bay Neutrino Experiment. This system consists of 24 identical units instrumented on 8 identical 20-ton liquid scintillator detectors. Each unit is a fully automated robotic system capable of deploying an LED and various radioactive sources into the detector along given vertical axes. Selected results from performance studies of the calibration system are reported

  18. Construction, Test And Calibration of the GLAST Silicon Tracker

    Energy Technology Data Exchange (ETDEWEB)

    Sgro, C.; Atwood, W.B.; Baldini, L.; Barbiellini, G.; Bellazzini, R.; Belli, F.; Bonamente, E.; Borden, T.; Bregeon, J.; Brez, A.; Brigida, M.; Caliandro, G.A.; Cecchi, C.; Cohen-Tanugi, J.; De Angelis, A.; Drell, P.; Favuzzi, C.; Fukazawa, Y.; Fusco, P.; Gargano, F.; Germani, S.; /INFN, Pisa /Pisa U. /UC, Santa Cruz /INFN, Trieste /Rome U.,Tor Vergata /SLAC /INFN, Bari /Bari U. /INFN, Perugia /Perugia U. /Udine U. /Hiroshima U. /Maryland U., JCA /Tokyo Inst. Tech. /JAXA, Sagamihara /INFN, Padua /Padua U. /Pisa, Scuola Normale Superiore /NASA, Goddard

    2009-06-05

    The Gamma-ray Large Area Space Telescope represents a great advance in space application of silicon detectors. With a surface of 80 m{sup 2} and about 1 M readout channels it is the largest silicon tracker ever built for a space experiment. GLAST is an astro-particle mission that will study the mostly unexplored, high energy (20 MeV-300 GeV) spectrum coming from active sources or diffused in the Universe. The detector integration and test phase is complete. The full instrument underwent environmental testing and the spacecraft integration phase has just started: the launch is foreseen in late 2007. In the meanwhile the spare modules are being used for instrument calibration and performance verification employing the CERN accelerator complex. A Calibration Unit has been exposed to photon, electron and hadron beams from a few GeV up to 300 GeV. We report on the status of the instrument and on the calibration campaign.

  19. Viscometry for liquids calibration of viscometers

    CERN Document Server

    Gupta, S V

    2014-01-01

    This book is written for scientists involved in the calibration of viscometers. A detailed description for stepping up procedures to establish the viscosity scale and obtaining sets of master viscometers is given in the book. Uncertainty considerations for standard oils of known viscosity are presented. The modern viscometers based on principles of  tuning fork, ultrasonic, PZT, plate waves, Love waves, micro-cantilever and vibration of optical fiber are discussed to inspire the reader to further research and to generate improved versions. The primary standard for viscosity is pure water. Measurements of its viscosity with accuracy/uncertainty achieved are described. The principles of rotational and oscillation viscometers are explained to enhance the knowledge in calibration work. Devices used for specific materials and viscosity in non SI units are discussed with respect to the need to correlate viscosity values obtained by various devices. The description of commercial viscometers meets the needs of the u...

  20. ORNL calibrations facility

    International Nuclear Information System (INIS)

    Berger, C.D.; Gupton, E.D.; Lane, B.H.; Miller, J.H.; Nichols, S.W.

    1982-08-01

    The ORNL Calibrations Facility is operated by the Instrumentation Group of the Industrial Safety and Applied Health Physics Division. Its primary purpose is to maintain radiation calibration standards for calibration of ORNL health physics instruments and personnel dosimeters. This report includes a discussion of the radioactive sources and ancillary equipment in use and a step-by-step procedure for calibration of those survey instruments and personnel dosimeters in routine use at ORNL

  1. Accurate KAP meter calibration as a prerequisite for optimisation in projection radiography

    International Nuclear Information System (INIS)

    Malusek, A.; Sandborg, M.; Alm Carlsson, G.

    2016-01-01

    Modern X-ray units register the air kerma-area product, PKA, with a built-in KAP meter. Some KAP meters show an energy-dependent bias comparable with the maximum uncertainty articulated by the IEC (25 %), adversely affecting dose-optimisation processes. To correct for the bias, a reference KAP meter calibrated at a standards laboratory and two calibration methods described here can be used to achieve an uncertainty of <7 % as recommended by IAEA. A computational model of the reference KAP meter is used to calculate beam quality correction factors for transfer of the calibration coefficient at the standards laboratory, Q 0 , to any beam quality, Q, in the clinic. Alternatively, beam quality corrections are measured with an energy-independent dosemeter via a reference beam quality in the clinic, Q 1 , to beam quality, Q. Biases up to 35 % of built-in KAP meter readings were noted. Energy-dependent calibration factors are needed for unbiased PKA. Accurate KAP meter calibration as a prerequisite for optimisation in projection radiography. (authors)

  2. SCIAMACHY Level 1 data: calibration concept and in-flight calibration

    Science.gov (United States)

    Lichtenberg, G.; Kleipool, Q.; Krijger, J. M.; van Soest, G.; van Hees, R.; Tilstra, L. G.; Acarreta, J. R.; Aben, I.; Ahlers, B.; Bovensmann, H.; Chance, K.; Gloudemans, A. M. S.; Hoogeveen, R. W. M.; Jongma, R. T. N.; Noël, S.; Piters, A.; Schrijver, H.; Schrijvers, C.; Sioris, C. E.; Skupin, J.; Slijkhuis, S.; Stammes, P.; Wuttke, M.

    2006-11-01

    The calibration of SCIAMACHY was thoroughly checked since the instrument was launched on-board ENVISAT in February 2002. While SCIAMACHY's functional performance is excellent since launch, a number of technical difficulties have appeared, that required adjustments to the calibration. The problems can be separated into three types: (1) Those caused by the instrument and/or platform environment. Among these are the high water content in the satellite structure and/or MLI layer. This results in the deposition of ice on the detectors in channels 7 and 8 which seriously affects the retrievals in the IR, mostly because of the continuous change of the slit function caused by scattering of the light through the ice layer. Additionally a light leak in channel 7 severely hampers any retrieval from this channel. (2) Problems due to errors in the on-ground calibration and/or data processing affecting for example the radiometric calibration. A new approach based on a mixture of on-ground and in-flight data is shortly described here. (3) Problems caused by principal limitations of the calibration concept, e.g. the possible appearance of spectral structures after the polarisation correction due to unavoidable errors in the determination of atmospheric polarisation. In this paper we give a complete overview of the calibration and problems that still have to be solved. We will also give an indication of the effect of calibration problems on retrievals where possible. Since the operational processing chain is currently being updated and no newly processed data are available at this point in time, for some calibration issues only a rough estimate of the effect on Level 2 products can be given. However, it is the intention of this paper to serve as a future reference for detailed studies into specific calibration issues.

  3. Calibration of working standard ionization chambers and dose standardization

    International Nuclear Information System (INIS)

    Abd Elmahoud, A. A. B.

    2011-01-01

    Measurements were performed for the calibration of two working standard ionization chambers in the secondary standard dosimetry laboratory of Sudan. 600 cc cylindrical former type and 1800 cc cylindrical radical radiation protection level ionization chambers were calibrated against 1000 cc spherical reference standard ionization chamber. The chamber were calibrated at X-ray narrow spectrum series with beam energies ranged from (33-116 KeV) in addition to 1''3''7''Cs beam with 662 KeV energy. The chambers 0.6 cc and 0.3 cc therapy level ionization were used for dose standardization and beam output calibrations of cobalt-60 radiotherapy machine located at the National Cancer Institute, University of Gazira. Concerning beam output measurements for 6''0''Co radiotherapy machine, dosimetric measurements were performed in accordance with the relevant per IAEA dosimetry protocols TRS-277 and TRS-398. The kinetic energy released per unit mass in air (air kerma) were obtained by multiplying the corrected electrometer reading (nC/min) by the calibration factors (Gy/n C) of the chambers from given in the calibration certificate. The uncertainty of measurements of air kerma were calculated for the all ionization chambers (combined uncertainty) the calibration factors of these ionization chambers then were calculated by comparing the reading of air kerma of secondary standard ionization chambers to than from radical and farmer chambers. The result of calibration working standard ionization chambers showed different calibration factors ranged from 0.99 to 1.52 for different radiation energies and these differences were due to chambers response and specification. The absorbed dose to to water calculated for therapy ionization chamber using two code of practice TRS-277 and TRS-398 as beam output for 6''0''Co radiotherapy machine and it can be used as a reference for future beam output calibration in radiotherapy dosimetry. The measurement of absorbed dose to water showed that the

  4. Regulatory audit to the Argentine teletherapy units

    International Nuclear Information System (INIS)

    Kay, J.M.

    1995-01-01

    As a consequence of an erroneous calibration of the ionization chamber dosimeter of a new licensed 6 Mev electron lineal accelerator that caused accidental overexposures in the first patients under treatment, the Regulatory Board decided to perform an audit to every teletherapy unit (accelerator and telegammatherapy units) in Argentina. The audit was carried out at the en of 1994 in 27 accelerators and 107 telegammatherapy units. The objectives of the audit were the control of the mechanical and optical aspects of the each accelerator and telegammatherapy unit and the control of the calibration of their radiation beams. The last action was undertaken to detect as soon as possible, any eventual accidental overexposure or underexposure of patients under treatment. The audit was conducted by the author and performed for a team of 22 inspectors who worked in groups of 2 to 3 people and completed the inspection of the teletherapy units of the whole country in 3 months. As a result of the audit, the Regulatory Board had in a short period a clear panorama of the performance of each teletherapy unit of the country and also confirmed that no accidental situation had happened. The audit revealed the necessity of imposing urgent regulatory requirements not related to the calibration of the radiation beam in 5% of the cases, as well as non-urgent requirements of different importance in approximately other 60% of the cases. (author)

  5. Calibration and verification of surface contamination meters --- Procedures and techniques

    International Nuclear Information System (INIS)

    Schuler, C; Butterweck, G.; Wernli, C.; Bochud, F.; Valley, J.-F.

    2007-03-01

    A standardised measurement procedure for surface contamination meters (SCM) is presented. The procedure aims at rendering surface contamination measurements to be simply and safely interpretable. Essential for the approach is the introduction and common use of the radionuclide specific quantity 'guideline value' specified in the Swiss Radiation Protection Ordinance as unit for the measurement of surface activity. The according radionuclide specific 'guideline value count rate' can be summarized as verification reference value for a group of radionuclides ('basis guideline value count rate'). The concept can be generalized for SCM of the same type or for SCM of different types using he same principle of detection. A SCM multi source calibration technique is applied for the determination of the instrument efficiency. Four different electron radiation energy regions, four different photon radiation energy regions and an alpha radiation energy region are represented by a set of calibration sources built according to ISO standard 8769-2. A guideline value count rate representing the activity per unit area of a surface contamination of one guideline value can be calculated for any radionuclide using instrument efficiency, radionuclide decay data, contamination source efficiency, guideline value averaging area (100 cm 2 ), and radionuclide specific guideline value. n this way, instrument responses for the evaluation of surface contaminations are obtained for radionuclides without available calibration sources as well as for short-Iived radionuclides, for which the continuous replacement of certified calibration sources can lead to unreasonable costs. SCM verification is based on surface emission rates of reference sources with an active area of 100 cm 2 . The verification for a given list of radionuclides is based on the radionuclide specific quantity guideline value count rate. Guideline value count rates for groups of radionuclides can be represented within the maximum

  6. Automated calibration system for a high-precision measurement of neutrino mixing angle θ{sub 13} with the Daya Bay antineutrino detectors

    Energy Technology Data Exchange (ETDEWEB)

    Liu, J., E-mail: jianglai.liu@sjtu.edu.cn [Department of Physics, Shanghai Jiao Tong University, Shanghai (China); Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Cai, B.; Carr, R. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Dwyer, D.A. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Lawrence Berkeley National Laboratory, Berkeley, CA (United States); Gu, W.Q.; Li, G.S. [Department of Physics, Shanghai Jiao Tong University, Shanghai (China); Qian, X. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Brookhaven National Laboratory, Upton, NY (United States); McKeown, R.D. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Department of Physics, College of William and Mary, Williamsburg, VA (United States); Tsang, R.H.M. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Wang, W. [Department of Physics, College of William and Mary, Williamsburg, VA (United States); Wu, F.F. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Zhang, C. [Kellogg Radiation Laboratory, California Institute of Technology, Pasadena, CA (United States); Brookhaven National Laboratory, Upton, NY (United States)

    2014-06-01

    We describe the automated calibration system for the antineutrino detectors in the Daya Bay Neutrino Experiment. This system consists of 24 identical units instrumented on 8 identical 20-ton liquid scintillator detectors. Each unit is a fully automated robotic system capable of deploying an LED and various radioactive sources into the detector along given vertical axes. Selected results from performance studies of the calibration system are reported.

  7. The needs for brachytherapy source calibrations in the United States

    International Nuclear Information System (INIS)

    Coursey, B.M.; Goodman, L.J.; Hoppes, D.D.; Loevinger, R.; McLaughlin, W.L.; Soares, C.G.; Weaver, J.T.

    1992-01-01

    Brachytherapy sources of beta and gamma radiation ('brachy' is from the Greek, meaning 'near') have a long history of use in interstitial, intracavitary, intraluminal, and ocular radiation therapy. In the past the US national standards for these sources were often specified in activity or milligram radium equivalent. With the introduction of new radionuclide sources to replace radium, source strength calibrations are now expressed as air kerma rate at a meter. In this paper, we review the NIST standards for brachytherapy sources, list some of the common radionuclides and source encapsulations in use in the US radiology community, and describe the latest NIST work, in collaboration with several US medical institutions, on a method of two- and three-dimensional dose mapping of brachytherapy sources using radiochromic films. (orig.)

  8. Calibration of the High Resolution Gamma Spectrometry Detection Systems from the Environmental Radioactivity Unit and Radiological Surveillance (ARU and RV); Calibracion de los Sistemas Detectores de Espectrometria Gamma de Alta Resolucion de la Unidad de Radiactividad Ambiental y Vigilancia Radiologica (URA y VR)

    Energy Technology Data Exchange (ETDEWEB)

    Suarez-Navarro, J. A.; Gasco, C.; Arana, M.; Suanez, A.

    2012-11-01

    In accordance with international regulations and with the quality criteria imposed by the internal quality system of the laboratory of gamma spectrometry, which belongs to the Unit of Environmental Radioactivity and Radiation Monitoring (URA y VR), it is necessary to perform a periodic calibration of the measurement systems of gamma spectrometry. The equipment calibration carries with the need for the preparation of the geometries of measurement and calibration in energy and in efficiencies. This report reflects the procedures for the preparation of the wide range of geometries available in the laboratory, as well as the different calibrations performed with those geometries in almost all the laboratory gamma detectors using the CANBERRA software Genie 2k. (Author) 9 refs.

  9. Spitzer/JWST Cross Calibration: IRAC Observations of Potential Calibrators for JWST

    Science.gov (United States)

    Carey, Sean J.; Gordon, Karl D.; Lowrance, Patrick; Ingalls, James G.; Glaccum, William J.; Grillmair, Carl J.; E Krick, Jessica; Laine, Seppo J.; Fazio, Giovanni G.; Hora, Joseph L.; Bohlin, Ralph

    2017-06-01

    We present observations at 3.6 and 4.5 microns using IRAC on the Spitzer Space Telescope of a set of main sequence A stars and white dwarfs that are potential calibrators across the JWST instrument suite. The stars range from brightnesses of 4.4 to 15 mag in K band. The calibration observations use a similar redundancy to the observing strategy for the IRAC primary calibrators (Reach et al. 2005) and the photometry is obtained using identical methods and instrumental photometric corrections as those applied to the IRAC primary calibrators (Carey et al. 2009). The resulting photometry is then compared to the predictions based on spectra from the CALSPEC Calibration Database (http://www.stsci.edu/hst/observatory/crds/calspec.html) and the IRAC bandpasses. These observations are part of an ongoing collaboration between IPAC and STScI investigating absolute calibration in the infrared.

  10. SDSS-IV/MaNGA: SPECTROPHOTOMETRIC CALIBRATION TECHNIQUE

    Energy Technology Data Exchange (ETDEWEB)

    Yan, Renbin; Sánchez-Gallego, José R. [Department of Physics and Astronomy, University of Kentucky, 505 Rose St., Lexington, KY 40506-0057 (United States); Tremonti, Christy; Bershady, Matthew A.; Eigenbrot, Arthur; Wake, David A. [Department of Astronomy, University of Winsconsin-Madison, 475 N. Charter Street, Madison, WI 53706-1582 (United States); Law, David R. [Space Telescope Science Institute, 3700 San Martin Drive, Baltimore, MD 21218 (United States); Schlegel, David J. [Physics Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94720-8160 (United States); Bundy, Kevin [Kavli IPMU (WPI), UTIAS, The University of Tokyo, Kashiwa, Chiba 277-8583 (Japan); Drory, Niv [McDonald Observatory, Department of Astronomy, University of Texas at Austin, 1 University Station, Austin, TX 78712-0259 (United States); MacDonald, Nicholas [Department of Astronomy, Box 351580, University of Washington, Seattle, WA 98195 (United States); Bizyaev, Dmitry [Apache Point Observatory, P.O. Box 59, sunspot, NM 88349 (United States); Blanc, Guillermo A. [Departamento de Astronomía, Universidad de Chile, Camino el Observatorio 1515, Las Condes, Santiago (Chile); Blanton, Michael R.; Hogg, David W. [Center for Cosmology and Particle Physics, Department of Physics, New York University, 4 Washington Place, New York, NY 10003 (United States); Cherinka, Brian [Dunlap Institute for Astronomy and Astrophysics, University of Toronto, 50 St. George Street, Toronto, Ontario M5S 3H4 (Canada); Gunn, James E. [Department of Astrophysical Sciences, Princeton University, Princeton, NJ 08544 (United States); Harding, Paul [Department of Astronomy, Case Western Reserve University, Cleveland, OH 44106 (United States); Sánchez, Sebastian F., E-mail: yanrenbin@uky.edu [Instituto de Astronomia, Universidad Nacional Autonoma de Mexico, A.P. 70-264, 04510 Mexico D.F. (Mexico); and others

    2016-01-15

    Mapping Nearby Galaxies at Apache Point Observatory (MaNGA), one of three core programs in the Sloan Digital Sky Survey-IV, is an integral-field spectroscopic survey of roughly 10,000 nearby galaxies. It employs dithered observations using 17 hexagonal bundles of 2″ fibers to obtain resolved spectroscopy over a wide wavelength range of 3600–10300 Å. To map the internal variations within each galaxy, we need to perform accurate spectral surface photometry, which is to calibrate the specific intensity at every spatial location sampled by each individual aperture element of the integral field unit. The calibration must correct only for the flux loss due to atmospheric throughput and the instrument response, but not for losses due to the finite geometry of the fiber aperture. This requires the use of standard star measurements to strictly separate these two flux loss factors (throughput versus geometry), a difficult challenge with standard single-fiber spectroscopy techniques due to various practical limitations. Therefore, we developed a technique for spectral surface photometry using multiple small fiber-bundles targeting standard stars simultaneously with galaxy observations. We discuss the principles of our approach and how they compare to previous efforts, and we demonstrate the precision and accuracy achieved. MaNGA's relative calibration between the wavelengths of Hα and Hβ has an rms of 1.7%, while that between [N ii] λ6583 and [O ii] λ3727 has an rms of 4.7%. Using extinction-corrected star formation rates and gas-phase metallicities as an illustration, this level of precision guarantees that flux calibration errors will be sub-dominant when estimating these quantities. The absolute calibration is better than 5% for more than 89% of MaNGA's wavelength range.

  11. SDSS-IV/MaNGA: SPECTROPHOTOMETRIC CALIBRATION TECHNIQUE

    International Nuclear Information System (INIS)

    Yan, Renbin; Sánchez-Gallego, José R.; Tremonti, Christy; Bershady, Matthew A.; Eigenbrot, Arthur; Wake, David A.; Law, David R.; Schlegel, David J.; Bundy, Kevin; Drory, Niv; MacDonald, Nicholas; Bizyaev, Dmitry; Blanc, Guillermo A.; Blanton, Michael R.; Hogg, David W.; Cherinka, Brian; Gunn, James E.; Harding, Paul; Sánchez, Sebastian F.

    2016-01-01

    Mapping Nearby Galaxies at Apache Point Observatory (MaNGA), one of three core programs in the Sloan Digital Sky Survey-IV, is an integral-field spectroscopic survey of roughly 10,000 nearby galaxies. It employs dithered observations using 17 hexagonal bundles of 2″ fibers to obtain resolved spectroscopy over a wide wavelength range of 3600–10300 Å. To map the internal variations within each galaxy, we need to perform accurate spectral surface photometry, which is to calibrate the specific intensity at every spatial location sampled by each individual aperture element of the integral field unit. The calibration must correct only for the flux loss due to atmospheric throughput and the instrument response, but not for losses due to the finite geometry of the fiber aperture. This requires the use of standard star measurements to strictly separate these two flux loss factors (throughput versus geometry), a difficult challenge with standard single-fiber spectroscopy techniques due to various practical limitations. Therefore, we developed a technique for spectral surface photometry using multiple small fiber-bundles targeting standard stars simultaneously with galaxy observations. We discuss the principles of our approach and how they compare to previous efforts, and we demonstrate the precision and accuracy achieved. MaNGA's relative calibration between the wavelengths of Hα and Hβ has an rms of 1.7%, while that between [N ii] λ6583 and [O ii] λ3727 has an rms of 4.7%. Using extinction-corrected star formation rates and gas-phase metallicities as an illustration, this level of precision guarantees that flux calibration errors will be sub-dominant when estimating these quantities. The absolute calibration is better than 5% for more than 89% of MaNGA's wavelength range

  12. Simultaneous calibration phantom commission and geometry calibration in cone beam CT

    Science.gov (United States)

    Xu, Yuan; Yang, Shuai; Ma, Jianhui; Li, Bin; Wu, Shuyu; Qi, Hongliang; Zhou, Linghong

    2017-09-01

    Geometry calibration is a vital step for describing the geometry of a cone beam computed tomography (CBCT) system and is a prerequisite for CBCT reconstruction. In current methods, calibration phantom commission and geometry calibration are divided into two independent tasks. Small errors in ball-bearing (BB) positioning in the phantom-making step will severely degrade the quality of phantom calibration. To solve this problem, we propose an integrated method to simultaneously realize geometry phantom commission and geometry calibration. Instead of assuming the accuracy of the geometry phantom, the integrated method considers BB centers in the phantom as an optimized parameter in the workflow. Specifically, an evaluation phantom and the corresponding evaluation contrast index are used to evaluate geometry artifacts for optimizing the BB coordinates in the geometry phantom. After utilizing particle swarm optimization, the CBCT geometry and BB coordinates in the geometry phantom are calibrated accurately and are then directly used for the next geometry calibration task in other CBCT systems. To evaluate the proposed method, both qualitative and quantitative studies were performed on simulated and realistic CBCT data. The spatial resolution of reconstructed images using dental CBCT can reach up to 15 line pair cm-1. The proposed method is also superior to the Wiesent method in experiments. This paper shows that the proposed method is attractive for simultaneous and accurate geometry phantom commission and geometry calibration.

  13. Multi-Site Calibration of Linear Reservoir Based Geomorphologic Rainfall-Runoff Models

    Directory of Open Access Journals (Sweden)

    Bahram Saeidifarzad

    2014-09-01

    Full Text Available Multi-site optimization of two adapted event-based geomorphologic rainfall-runoff models was presented using Non-dominated Sorting Genetic Algorithm (NSGA-II method for the South Fork Eel River watershed, California. The first model was developed based on Unequal Cascade of Reservoirs (UECR and the second model was presented as a modified version of Geomorphological Unit Hydrograph based on Nash’s model (GUHN. Two calibration strategies were considered as semi-lumped and semi-distributed for imposing (or unimposing the geomorphology relations in the models. The results of models were compared with Nash’s model. Obtained results using the observed data of two stations in the multi-site optimization framework showed reasonable efficiency values in both the calibration and the verification steps. The outcomes also showed that semi-distributed calibration of the modified GUHN model slightly outperformed other models in both upstream and downstream stations during calibration. Both calibration strategies for the developed UECR model during the verification phase showed slightly better performance in the downstream station, but in the upstream station, the modified GUHN model in the semi-lumped strategy slightly outperformed the other models. The semi-lumped calibration strategy could lead to logical lag time parameters related to the basin geomorphology and may be more suitable for data-based statistical analyses of the rainfall-runoff process.

  14. SPRT Calibration Uncertainties and Internal Quality Control at a Commercial SPRT Calibration Facility

    Science.gov (United States)

    Wiandt, T. J.

    2008-06-01

    The Hart Scientific Division of the Fluke Corporation operates two accredited standard platinum resistance thermometer (SPRT) calibration facilities, one at the Hart Scientific factory in Utah, USA, and the other at a service facility in Norwich, UK. The US facility is accredited through National Voluntary Laboratory Accreditation Program (NVLAP), and the UK facility is accredited through UKAS. Both provide SPRT calibrations using similar equipment and procedures, and at similar levels of uncertainty. These uncertainties are among the lowest available commercially. To achieve and maintain low uncertainties, it is required that the calibration procedures be thorough and optimized. However, to minimize customer downtime, it is also important that the instruments be calibrated in a timely manner and returned to the customer. Consequently, subjecting the instrument to repeated calibrations or extensive repeated measurements is not a viable approach. Additionally, these laboratories provide SPRT calibration services involving a wide variety of SPRT designs. These designs behave differently, yet predictably, when subjected to calibration measurements. To this end, an evaluation strategy involving both statistical process control and internal consistency measures is utilized to provide confidence in both the instrument calibration and the calibration process. This article describes the calibration facilities, procedure, uncertainty analysis, and internal quality assurance measures employed in the calibration of SPRTs. Data will be reviewed and generalities will be presented. Finally, challenges and considerations for future improvements will be discussed.

  15. Study on low-cost calibration-free pH sensing with disposable optical sensors

    International Nuclear Information System (INIS)

    Ge Xudong; Kostov, Yordan; Tolosa, Leah; Rao, Govind

    2012-01-01

    Highlights: ► Development of disposable calibration-free optical pH sensor. ► Derivation of theoretical calibration model. ► Strategy to eliminate inter-coaster inconsistency. - Abstract: As labor costs become more expensive, less labor-intensive disposable devices have become more ubiquitous. Similarly, the disposable optical pH sensor developed in our lab could provide a convenient yet cost-effective way for pH sensing in processes that require stringent pH control. This optical pH sensor is prepared in uniform individual lots of 100–200 sensors per lot. Calibration is accomplished on a few randomly selected sensors out of each lot. We show that all others in the same lot can then be used directly without requiring individual calibration. In this paper, a calibration model is derived to include all the factors that affect the signal of the disposable sensor. Experimental results show that the derived calibration model fits the experimental data. The readings of 28 randomly selected disposable sensors with 4 sensors from each of the 7 lots show an error less than 0.1 pH units in the useful sensing range of the sensor. The calibration model indicates that if further improvement on precision is desired, more uniform porous material and more advanced coating techniques will be required. When it comes to the effects of the varying coasters, house-made low-cost fluorometers, the variability in the brightness ratio of the blue-to-violet LEDs is the primary reason for the lack of precision. Other factors like LED light intensity distribution, optical properties of the filters and electronics also contribute to the coaster-to-coaster difference, but to a lesser extent. Two different methods for correcting the instrument variations were introduced. After correction, the collective reading errors for all the tested instruments were reduced to less than 0.2 pH units within the sensor's useful sensing range. Based on this result, our lab is currently implementing

  16. A Monte Carlo modeling alternative for the API Gamma Ray Calibration Facility.

    Science.gov (United States)

    Galford, J E

    2017-04-01

    The gamma ray pit at the API Calibration Facility, located on the University of Houston campus, defines the API unit for natural gamma ray logs used throughout the petroleum logging industry. Future use of the facility is uncertain. An alternative method is proposed to preserve the gamma ray API unit definition as an industry standard by using Monte Carlo modeling to obtain accurate counting rate-to-API unit conversion factors for gross-counting and spectral gamma ray tool designs. Copyright © 2017 Elsevier Ltd. All rights reserved.

  17. Evaluating Statistical Process Control (SPC) techniques and computing the uncertainty of force calibrations

    Science.gov (United States)

    Navard, Sharon E.

    1989-01-01

    In recent years there has been a push within NASA to use statistical techniques to improve the quality of production. Two areas where statistics are used are in establishing product and process quality control of flight hardware and in evaluating the uncertainty of calibration of instruments. The Flight Systems Quality Engineering branch is responsible for developing and assuring the quality of all flight hardware; the statistical process control methods employed are reviewed and evaluated. The Measurement Standards and Calibration Laboratory performs the calibration of all instruments used on-site at JSC as well as those used by all off-site contractors. These calibrations must be performed in such a way as to be traceable to national standards maintained by the National Institute of Standards and Technology, and they must meet a four-to-one ratio of the instrument specifications to calibrating standard uncertainty. In some instances this ratio is not met, and in these cases it is desirable to compute the exact uncertainty of the calibration and determine ways of reducing it. A particular example where this problem is encountered is with a machine which does automatic calibrations of force. The process of force calibration using the United Force Machine is described in detail. The sources of error are identified and quantified when possible. Suggestions for improvement are made.

  18. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities

    International Nuclear Information System (INIS)

    Costa, Alessandro Martins da

    1999-01-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  19. A primary standard for the calibration of sniffer test leak devices

    Science.gov (United States)

    Jousten, Karl; Becker, Ute

    2009-10-01

    Test leaks with a gas flow to atmospheric pressure are often called sniffer test leaks. They are used to calibrate leak detectors for sniffing applications. Sniffer test leaks need calibration against a standard. A primary standard for the calibration of sniffer test leaks with relatively low measurement uncertainties is described. It is assured that the measurement result is traceable to the relevant SI units and that there is a well-known and complete measurement uncertainty budget. The measurement range of the system is from 4 × 10-11 mol s-1 (corresponding to 10-4 Pa l s-1 at 23 °C) to 4 × 10-9 mol s-1 (10-2 Pa l s-1 at 23 °C), which is the most often needed range in industry of around 1 g loss per year of the cooling agent R134a. The temperature where the calibration can be carried out may vary from 18 °C to 30 °C. The flow rate of any test gas not condensing in this temperature range can be measured.

  20. Absolute calibration of TFTR neutron detectors for D-T plasma operation

    International Nuclear Information System (INIS)

    Jassby, D.L.; Johnson, L.C.; Roquemore, A.L.; Strachan, J.D.; Johnson, D.W.; Medley, S.S.; Young, K.M.

    1995-03-01

    The two most sensitive TFTR fission-chamber detectors were absolutely calibrated in situ by a D-T neutron generator (∼5 x 10 7 n/s) rotated once around the torus in each direction, with data taken at about 45 positions. The combined uncertainty for determining fusion neutron rates, including the uncertainty in the total neutron generator output (±9%), counting statistics, the effect of coil coolant, detector stability, cross-calibration to the current mode or log Campbell mode and to other fission chambers, and plasma position variation, is about ±13%. The NE-451 (ZnS) scintillators and 4 He proportional counters that view the plasma in up to 10 collimated sightlines were calibrated by scanning. the neutron generator radially and toroidally in the horizontal midplane across the flight tubes of 7 cm diameter. Spatial integration of the detector responses using the calibrated signal per unit chord-integrated neutron emission gives the global neutron source strength with an overall uncertainty of ±14% for the scintillators and ±15% for the 4 He counters

  1. In flight calibrations of Ibis/PICsIT

    International Nuclear Information System (INIS)

    Malaguti, G.; Di Cocco, G.; Foschini, L.; Stephen, J.B.; Bazzano, A.; Ubertini, P.; Bird, A.J.; Laurent, P.; Segreto, A.

    2003-01-01

    PICsIT (Pixellated Imaging Caesium Iodide Telescope) is the high energy detector of the IBIS telescope on-board the INTEGRAL satellite. It consists of 4096 independent detection units, ∼ 0.7 cm 2 in cross-section, operating in the energy range between 175 keV and 10 MeV. The intrinsically low signal to noise ratio in the gamma-ray astronomy domain implies very long observations, lasting 10 5 - 10 6 s. Moreover, the image formation principle on which PICsIT works is that of coded imaging in which the entire detection plane contributes to each decoded sky pixel. For these two main reasons, the monitoring, and possible correction, of the spatial and temporal non-uniformity of pixel performances, especially in terms of gain and energy resolution, is of paramount importance. The IBIS on-board 22 Na calibration source allows the calibration of each pixel at an accuracy of <0.5% by integrating the data from a few revolutions at constant temperature. The two calibration lines, at 511 and 1275 keV, allow also the measurement and monitoring of the PICsIT energy resolution which proves to be very stable at ∼ 19% and ∼ 9% (FWHM) respectively, and consistent with the values expected analytical predictions checked against pre-launch tests. (authors)

  2. Rotating-coil calibration in a reference quadrupole, considering roll-angle misalignment and higher-order harmonics

    CERN Document Server

    AUTHOR|(CDS)2075492; Buzio, Marco; Köster, Oliver; Russenschuck, Stephan; Severino, Giordana

    2016-01-01

    A method is proposed for calibrating the radius of a rotating coil sensor by relaxing the metrological constraints on alignment and field errors of the reference quadrupole. A coil radius calibration considering a roll-angle misalignment of the measurement bench, the magnet, and the motor-drive unit is analyzed. Then, the error arising from higher-order harmonic field imperfections in the reference quadrupole is assessed. The method is validated by numerical field computation for both the higher-order harmonic errors and the roll-angle misalignment. Finally, an experimental proof-of-principle demonstration is car-ried out in a calibration magnet with sextupole harmonic.

  3. Scanner calibration revisited

    Directory of Open Access Journals (Sweden)

    Pozhitkov Alexander E

    2010-07-01

    Full Text Available Abstract Background Calibration of a microarray scanner is critical for accurate interpretation of microarray results. Shi et al. (BMC Bioinformatics, 2005, 6, Art. No. S11 Suppl. 2. reported usage of a Full Moon BioSystems slide for calibration. Inspired by the Shi et al. work, we have calibrated microarray scanners in our previous research. We were puzzled however, that most of the signal intensities from a biological sample fell below the sensitivity threshold level determined by the calibration slide. This conundrum led us to re-investigate the quality of calibration provided by the Full Moon BioSystems slide as well as the accuracy of the analysis performed by Shi et al. Methods Signal intensities were recorded on three different microarray scanners at various photomultiplier gain levels using the same calibration slide from Full Moon BioSystems. Data analysis was conducted on raw signal intensities without normalization or transformation of any kind. Weighted least-squares method was used to fit the data. Results We found that initial analysis performed by Shi et al. did not take into account autofluorescence of the Full Moon BioSystems slide, which led to a grossly distorted microarray scanner response. Our analysis revealed that a power-law function, which is explicitly accounting for the slide autofluorescence, perfectly described a relationship between signal intensities and fluorophore quantities. Conclusions Microarray scanners respond in a much less distorted fashion than was reported by Shi et al. Full Moon BioSystems calibration slides are inadequate for performing calibration. We recommend against using these slides.

  4. New blackbody standard for the evaluation and calibration of tympanic ear thermometers at the NPL, United Kingdom

    Science.gov (United States)

    McEvoy, Helen C.; Simpson, Robert; Machin, Graham

    2004-04-01

    The use of infrared tympanic thermometers for monitoring patient health is widespread. However, studies into the performance of these thermometers have questioned their accuracy and repeatability. To give users confidence in these devices, and to provide credibility in the measurements, it is necessary for them to be tested using an accredited, standard blackbody source, with a calibration traceable to the International Temperature Scale of 1990 (ITS-90). To address this need the National Physical Laboratory (NPL), UK, has recently set up a primary ear thermometer calibration (PET-C) source for the evaluation and calibration of tympanic (ear) thermometers over the range from 15 °C to 45 °C. The overall uncertainty of the PET-C source is estimated to be +/- 0.04 °C at k = 2. The PET-C source meets the requirements of the European Standard EN 12470-5: 2003 Clinical thermometers. It consists of a high emissivity blackbody cavity immersed in a bath of stirred liquid. The temperature of the blackbody is determined using an ITS-90 calibrated platinum resistance thermometer inserted close to the rear of the cavity. The temperature stability and uniformity of the PET-C source was evaluated and its performance validated. This paper provides a description of the PET-C along with the results of the validation measurements. To further confirm the performance of the PET-C source it was compared to the standard ear thermometer calibration sources of the National Metrology Institute of Japan (NMIJ), Japan and the Physikalisch-Technische Bundesanstalt (PTB), Germany. The results of this comparison will also be briefly discussed. The PET-C source extends the capability for testing ear thermometers offered by the NPL body temperature fixed-point source, described previously. An update on the progress with the commercialisation of the fixed-point source will be given.

  5. An accurate calibration method for accelerometer nonlinear scale factor on a low-cost three-axis turntable

    International Nuclear Information System (INIS)

    Pan, Jianye; Zhang, Chunxi; Cai, Qingzhong

    2014-01-01

    Strapdown inertial navigation system (SINS) requirements are very demanding on gyroscopes and accelerometers as well as on calibration. To improve the accuracy of SINS, high-accuracy calibration is needed. Adding the accelerometer nonlinear scale factor into the model and reducing estimation errors is essential for improving calibration methods. In this paper, the inertial navigation error model is simplified, including only velocity and tilt errors. Based on the simplified error model, the relationship between the navigation errors (the rates of change of velocity errors) and the inertial measurement unit (IMU) calibration parameters is presented. A tracking model is designed to estimate the rates of change of velocity errors. With a special calibration procedure consisting of six rotation sequences, the accelerometer nonlinear scale factor errors can be computed by the estimates of the rates of change of velocity errors. Simulation and laboratory test results show that the accelerometer nonlinear scale factor can be calibrated with satisfactory accuracy on a low-cost three-axis turntable in several minutes. The comparison with the traditional calibration method highlights the superior performance of the proposed calibration method without precise orientation control. In addition, the proposed calibration method saves a lot of time in comparison with the multi-position calibration method. (paper)

  6. History, organization, and oversight of the accredited dosimetry calibration laboratories by the AAPM

    Energy Technology Data Exchange (ETDEWEB)

    Rozenfeld, M. [St. James Hospital and Health Centers, Chicago Heights, IL (United States)

    1993-12-31

    For more than 20 years, the American Association of Physicists in Medicine (AAPM) has operated an accreditation program for secondary standards laboratories that calibrate radiation measuring instruments. Except for one short period, that program has been able to provide the facilities to satisfy the national need for accurate calibrations of such instruments. That exception, in 1981, due to the combination of the U.S. Nuclear Regulatory Commission (NRC) requiring instrument calibrations by users of cobalt-60 teletherapy units and the withdrawal of one of the three laboratories accredited at that time. However, after successful operation as a Task Group of the Radiation Therapy Committee (RTC) of the AAPM for two decades, a reorganization of this structure is now under serious consideration by the administration of the AAPM.

  7. History, organization, and oversight of the accredited dosimetry calibration laboratories by the AAPM

    International Nuclear Information System (INIS)

    Rozenfeld, M.

    1993-01-01

    For more than 20 years, the American Association of Physicists in Medicine (AAPM) has operated an accreditation program for secondary standards laboratories that calibrate radiation measuring instruments. Except for one short period, that program has been able to provide the facilities to satisfy the national need for accurate calibrations of such instruments. That exception, in 1981, due to the combination of the U.S. Nuclear Regulatory Commission (NRC) requiring instrument calibrations by users of cobalt-60 teletherapy units and the withdrawal of one of the three laboratories accredited at that time. However, after successful operation as a Task Group of the Radiation Therapy Committee (RTC) of the AAPM for two decades, a reorganization of this structure is now under serious consideration by the administration of the AAPM

  8. Intersatellite Calibration of Microwave Radiometers for GPM

    Science.gov (United States)

    Wilheit, T. T.

    2010-12-01

    The aim of the GPM mission is to measure precipitation globally with high temporal resolution by using a constellation of satellites logically united by the GPM Core Satellite which will be in a non-sunsynchronous, medium inclination orbit. The usefulness of the combined product depends on the consistency of precipitation retrievals from the various microwave radiometers. The calibration requirements for this consistency are quite daunting requiring a multi-layered approach. The radiometers can vary considerably in their frequencies, view angles, polarizations and spatial resolutions depending on their primary application and other constraints. The planned parametric algorithms will correct for the varying viewing parameters, but they are still vulnerable to calibration errors, both relative and absolute. The GPM Intersatellite Calibration Working Group (aka X-CAL) will adjust the calibration of all the radiometers to a common consensus standard for the GPM Level 1C product to be used in precipitation retrievals. Finally, each Precipitation Algorithm Working Group must have its own strategy for removing the residual errors. If the final adjustments are small, the credibility of the precipitation retrievals will be enhanced. Before intercomparing, the radiometers must be self consistent on a scan-wise and orbit-wise basis. Pre-screening for this consistency constitutes the first step in the intercomparison. The radiometers are then compared pair-wise with the microwave radiometer (GMI) on the GPM Core Satellite. Two distinct approaches are used for sake of cross-checking the results. On the one hand, nearly simultaneous observations are collected at the cross-over points of the orbits and the observations of one are converted to virtual observations of the other using a radiative transfer model to permit comparisons. The complementary approach collects histograms of brightness temperature from each instrument. In each case a model is needed to translate the

  9. Design of a Two-Step Calibration Method of Kinematic Parameters for Serial Robots

    Science.gov (United States)

    WANG, Wei; WANG, Lei; YUN, Chao

    2017-03-01

    Serial robots are used to handle workpieces with large dimensions, and calibrating kinematic parameters is one of the most efficient ways to upgrade their accuracy. Many models are set up to investigate how many kinematic parameters can be identified to meet the minimal principle, but the base frame and the kinematic parameter are indistinctly calibrated in a one-step way. A two-step method of calibrating kinematic parameters is proposed to improve the accuracy of the robot's base frame and kinematic parameters. The forward kinematics described with respect to the measuring coordinate frame are established based on the product-of-exponential (POE) formula. In the first step the robot's base coordinate frame is calibrated by the unit quaternion form. The errors of both the robot's reference configuration and the base coordinate frame's pose are equivalently transformed to the zero-position errors of the robot's joints. The simplified model of the robot's positioning error is established in second-power explicit expressions. Then the identification model is finished by the least square method, requiring measuring position coordinates only. The complete subtasks of calibrating the robot's 39 kinematic parameters are finished in the second step. It's proved by a group of calibration experiments that by the proposed two-step calibration method the average absolute accuracy of industrial robots is updated to 0.23 mm. This paper presents that the robot's base frame should be calibrated before its kinematic parameters in order to upgrade its absolute positioning accuracy.

  10. Calibration-on-the-spot”: How to calibrate an EMCCD camera from its images

    DEFF Research Database (Denmark)

    Mortensen, Kim; Flyvbjerg, Henrik

    2016-01-01

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction-......-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information....

  11. Calibration of reference KAP-meters at SSDL and cross calibration of clinical KAP-meters

    International Nuclear Information System (INIS)

    Hetland, Per O.; Friberg, Eva G.; Oevreboe, Kirsti M.; Bjerke, Hans H.

    2009-01-01

    In the summer of 2007 the secondary standard dosimetry laboratory (SSDL) in Norway established a calibration service for reference air-kerma product meter (KAP-meter). The air-kerma area product, PKA, is a dosimetric quantity that can be directly related to the patient dose and used for risk assessment associated with different x-ray examinations. The calibration of reference KAP-meters at the SSDL gives important information on parameters influencing the calibration factor for different types of KAP-meters. The use of reference KAP-meters calibrated at the SSDL is an easy and reliable way to calibrate or verify the PKA indicated by the x-ray equipment out in the clinics. Material and methods. Twelve KAP-meters were calibrated at the SSDL by use of the substitution method at five diagnostic radiation qualities (RQRs). Results. The calibration factors varied from 0.94 to 1.18. The energy response of the individual KAP-meters varied by a total of 20% between the different RQRs and the typical chamber transmission factors ranged from 0.78 to 0.91. Discussion. It is important to use a calibrated reference KAP-meter and a harmonised calibration method in the PKA calibration in hospitals. The obtained uncertainty in the PKA readings is comparable with other calibration methods if the information in the calibration certificate is correct used, corrections are made and proper positioning of the KAP-chamber is performed. This will ensure a reliable estimate of the patient dose and a proper optimisation of conventional x-ray examinations and interventional procedures

  12. Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit.

    Science.gov (United States)

    Najafi, Mohsen; Geraily, Ghazale; Shirazi, Alireza; Esfahani, Mahbod; Teimouri, Javad

    2017-01-01

    In recent years, Gafchromic films are used as an advanced instrument for dosimetry systems. The EBT3 films are a new generation of Gafchromic films. Our main interest is to compare the response of the EBT3 films exposed to gamma rays provided by the Theratron 780C as a conventional radiotherapy system and the Leksell Gamma Knife as a stereotactic radiotherapy system (SRS). Both systems use Cobalt-60 sources, thus using the same energy. However, other factors such as source-to-axis distance, number of sources, dose rate, direction of irradiation, shape of phantom, the field shape of radiation, and different scatter contribution may influence the calibration curve. Calibration curves for the 2 systems were measured and plotted for doses ranging from 0 to 40 Gy at the red and green channels. The best fitting curve was obtained with the Levenberg-Marquardt algorithm. Also, the component of dose uncertainty was obtained for any calibration curve. With the best fitting curve for the EBT3 films, we can use the calibration curve to measure the absolute dose in radiation therapy. Although there is a small deviation between the 2 curves, the p-value at any channel shows no significant difference between the 2 calibration curves. Therefore, the calibration curve for each system can be the same because of minor differences. The results show that with the best fitting curve from measured data, while considering the measurement uncertainties related to them, the EBT3 calibration curve can be used to measure the unknown dose both in SRS and in conventional radiotherapy. Copyright © 2017. Published by Elsevier Inc.

  13. A novel single-step, multipoint calibration method for instrumented Lab-on-Chip systems

    DEFF Research Database (Denmark)

    Pfreundt, Andrea; Patou, François; Zulfiqar, Azeem

    2014-01-01

    for instrument-based PoC blood biomarker analysis systems. Motivated by the complexity of associating high-accuracy biosensing using silicon nanowire field effect transistors with ease of use for the PoC system user, we propose a novel one-step, multipoint calibration method for LoC-based systems. Our approach...... specifically addresses the important interfaces between a novel microfluidic unit to integrate the sensor array and a mobile-device hardware accessory. A multi-point calibration curve is obtained by generating a defined set of reference concentrations from a single input. By consecutively splitting the flow...

  14. Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility

    Science.gov (United States)

    Mahavarkar, Prasanna; John, Jacob; Dhapre, Vijay; Dongre, Varun; Labde, Sachin

    2018-04-01

    A tri-axial square Helmholtz coil system for the study of palaeomagnetic studies, manufactured by GEOFYZIKA (former Czechoslovakia), was successfully commissioned at the Alibag Magnetic Observatory (IAGA code: ABG) in the year 1985. This system was used for a few years, after which the system encountered technical problems with the control unit. Rectification of the unit could not be undertaken, as the information document related to this system was not available, and as a result the system had been lying in an unused state for a long time, until 2015, when the system was recommissioned and upgraded to a test facility for calibrating the magnetometer sensors. We have upgraded the system with a constant current source and a data-logging unit. Both of these units have been designed and developed in the institute laboratory. Also, re-measurements of the existing system have been made thoroughly. The upgraded system is semi-automatic, enabling non-specialists to operate it after a brief period of instruction. This facility is now widely used at the parent institute and external institutions to calibrate magnetometers and it also serves as a national facility. Here the design of this system with the calibration results for the space-borne fluxgate magnetometers is presented.

  15. Tri-axial square Helmholtz coil system at the Alibag Magnetic Observatory: upgraded to a magnetic sensor calibration facility

    Directory of Open Access Journals (Sweden)

    P. Mahavarkar

    2018-04-01

    Full Text Available A tri-axial square Helmholtz coil system for the study of palaeomagnetic studies, manufactured by GEOFYZIKA (former Czechoslovakia, was successfully commissioned at the Alibag Magnetic Observatory (IAGA code: ABG in the year 1985. This system was used for a few years, after which the system encountered technical problems with the control unit. Rectification of the unit could not be undertaken, as the information document related to this system was not available, and as a result the system had been lying in an unused state for a long time, until 2015, when the system was recommissioned and upgraded to a test facility for calibrating the magnetometer sensors. We have upgraded the system with a constant current source and a data-logging unit. Both of these units have been designed and developed in the institute laboratory. Also, re-measurements of the existing system have been made thoroughly. The upgraded system is semi-automatic, enabling non-specialists to operate it after a brief period of instruction. This facility is now widely used at the parent institute and external institutions to calibrate magnetometers and it also serves as a national facility. Here the design of this system with the calibration results for the space-borne fluxgate magnetometers is presented.

  16. Investigation on calibration parameter of mammography calibration facilities at MINT

    International Nuclear Information System (INIS)

    Asmaliza Hashim; Wan Hazlinda Ismail; Md Saion Salikin; Muhammad Jamal Md Isa; Azuhar Ripin; Norriza Mohd Isa

    2004-01-01

    A mammography calibration facility has been established in the Medical Physics Laboratory, Malaysian Institute for Nuclear Technology Research (MINT). The calibration facility is established at the national level mainly to provide calibration services for radiation measuring test instruments or test tools used in quality assurance programme in mammography, which is being implemented in Malaysia. One of the accepted parameters that determine the quality of a radiation beam is the homogeneity coefficient. It is determined from the values of the 1 st and 2 nd Half Value Layer (HVL). In this paper, the consistency of the mammography machine beam qualities that is available in MINT, is investigated and presented. For calibration purposes, five radiation qualities namely 23, 25, 28, 30 and 35 kV, selectable from the control panel of the X-ray machine is used. Important parameters that are set for this calibration facility are exposure time, tube current, focal spot to detector distance (FDD) and beam size at specific distance. The values of homogeneity coefficient of this laboratory for the past few years tip to now be presented in this paper. Backscatter radiations are also considered in this investigation. (Author)

  17. Calibration Under Uncertainty.

    Energy Technology Data Exchange (ETDEWEB)

    Swiler, Laura Painton; Trucano, Timothy Guy

    2005-03-01

    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  18. Analysis of thermal power calibration method

    International Nuclear Information System (INIS)

    Zagar, T.; Ravnik, M.; Persic, A.

    2000-01-01

    The methods for determining fuel element burnup have recently become interesting because of activities related to the shipment of highly enriched fuel elements back to the United States for final disposal before 2009. The most common and practical method for determining fuel element burnup in research reactors is reactor calculation. Experience has shown that burnup calculations become complicated and biased with uncertainties if a long period of reactor operation must be reproduced. Besides this, accuracy of calculated burnup is always limited with accuracy of reactor power calibration, since burnup calculation is based on calculated power density distribution, which is usually expressed in terms of power released per fuel element and normalised to the reactor power It is obvious that reactor thermal power calibration is very important for precise fuel element burnup calculation. Calculated fuel element burnup is linearly dependent on the thermal reactor power. The reactor power level may be determined from measured absolute thermal flux distribution across the core in the horizontal and vertical planes. Flux distributions are measured with activation of cadmium covered and bare foils irradiated by the steady reactor power. But it should be realised that this method is time consuming and not accurate. This method is practical only for zero power reactors and is in practice very seldom performed for other reactors (e.g. for TRIGA reactor in Ljubljana absolute thermal flux distribution was not performed since reactor reconstruction in 1991). In case of power reactors and research reactors in which a temperature rise across the core is produced and measured than a heat balance method is the most common and accurate method of determining the power output of the core. The purpose of this paper is to analyse the accuracy of calorimetric reactor power calibration method and to analyse the influence of control rod position on nuclear detector reading for TRIGA reactors

  19. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten

    2007-01-01

    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve...... the statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited...

  20. Calibration of Flick standards

    International Nuclear Information System (INIS)

    Thalmann, Ruedi; Spiller, Jürg; Küng, Alain; Jusko, Otto

    2012-01-01

    Flick standards or magnification standards are widely used for an efficient and functional calibration of the sensitivity of form measuring instruments. The results of a recent measurement comparison have shown to be partially unsatisfactory and revealed problems related to the calibration of these standards. In this paper the influence factors for the calibration of Flick standards using roundness measurement instruments are discussed in detail, in particular the bandwidth of the measurement chain, residual form errors of the device under test, profile distortions due to the diameter of the probing element and questions related to the definition of the measurand. The different contributions are estimated using simulations and are experimentally verified. Also alternative methods to calibrate Flick standards are investigated. Finally the practical limitations of Flick standard calibration are shown and the usability of Flick standards both to calibrate the sensitivity of roundness instruments and to check the filter function of such instruments is analysed. (paper)

  1. Optical calibration and test of the VLT Deformable Secondary Mirror

    Science.gov (United States)

    Briguglio, Runa; Xompero, Marco; Riccardi, Armando; Andrighettoni, Mario; Pescoller, Dietrich; Biasi, Roberto; Gallieni, Daniele; Vernet, Elise; Kolb, Johann; Arsenault, Robin; Madec, Pierre-Yves

    2013-12-01

    The Deformable Secondary Mirror (DSM) for the VLT (ESO) represents the state-of-art of the large-format deformable mirror technology with its 1170 voice-coil actuators and its internal metrology based on actuator co-located capacitive sensors to control the shape of the 1.12m-diameter 2mm-thick convex shell. The present paper reports the results of the optical characterization of the mirror unit with the ASSIST facility located at ESO-Garching and executed in a collaborative effort by ESO, INAF-Osservatorio Astrofisico di Arcetri and the DSM manufacturing companies (Microgate s.r.l. and A.D.S. International s.r.l.). The main purposes of the tests are the optical characterization of the shell flattening residuals, the corresponding calibration of flattening commands, the optical calibration of the capacitive sensors and the optical calibration of the mirror influence functions. The results are used for the optical acceptance of the DSM and to allow the next test phase coupling the DSM with the wave-front sensor modules of the new Adaptive Optics Facility (AOF) of ESO.

  2. Synthesis Polarimetry Calibration

    Science.gov (United States)

    Moellenbrock, George

    2017-10-01

    Synthesis instrumental polarization calibration fundamentals for both linear (ALMA) and circular (EVLA) feed bases are reviewed, with special attention to the calibration heuristics supported in CASA. Practical problems affecting modern instruments are also discussed.

  3. Sandia WIPP calibration traceability

    Energy Technology Data Exchange (ETDEWEB)

    Schuhen, M.D. [Sandia National Labs., Albuquerque, NM (United States); Dean, T.A. [RE/SPEC, Inc., Albuquerque, NM (United States)

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities.

  4. Sandia WIPP calibration traceability

    International Nuclear Information System (INIS)

    Schuhen, M.D.; Dean, T.A.

    1996-05-01

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities

  5. Program for large scanning-measuring table calibration on-line with the TRA-1001 computer

    International Nuclear Information System (INIS)

    Kalmykova, L.P.; Ososkov, G.A.; Pogodina, G.A.; Skryl', I.I.

    1975-01-01

    A description of a programme variant for complex calibration of the system of boards BPS-2, is given in which the stage of measurements and accumulation of calibration data is separated from that of data processing on CDC-1604A computers. Stage-by-stage calibration allows for simultaneous and parallel measurements on all the 6 boards. Such a possibility of the boards' operation on a computer-controlled line has been used as checking of the operators' actions with current control of the data obtained. Mathematical formulae are followed by programmes' descriptions. Reception and accumulation of coordinates of the units of the calibrating plate, with a highly precise net of straight lines, are accomplished with the help of the DATREC programme working on a guiding TPA-1001 computer with simultaneous calibration of all the six BPS-2 boards. The DATREC programme is written in the SLANG-1 language. The CALBPS programme works on CDC-1604A computer, is written in FORTRAN, and calculates transformation coefficients and corresponding precision characteristics. The work has resulted in reducing the calibration time from 10-15 min. per board to 10-15 min. per all the 6 boards. The time of accumulation in the memory TPA-1001 and of recording on the CDC-608 tape recorder of calibration data is from 3 to 5 min.; the remaining time is spent on processing on a bigger CDC-1604A computer. Examples of typical output and certain results of calibration measurements are given

  6. 21 CFR 870.4300 - Cardiopulmonary bypass gas control unit.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Cardiopulmonary bypass gas control unit. 870.4300... bypass gas control unit. (a) Identification. A cardiopulmonary bypass gas control unit is a device used to control and measure the flow of gas into the oxygenator. The device is calibrated for a specific...

  7. Standard practice of calibration of force-measuring instruments for verifying the force indication of testing machines

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2006-01-01

    1.1 The purpose of this practice is to specify procedures for the calibration of force-measuring instruments. Procedures are included for the following types of instruments: 1.1.1 Elastic force-measuring instruments, and 1.1.2 Force-multiplying systems, such as balances and small platform scales. Note 1Verification by deadweight loading is also an acceptable method of verifying the force indication of a testing machine. Tolerances for weights for this purpose are given in Practices E 4; methods for calibration of the weights are given in NIST Technical Note 577, Methods of Calibrating Weights for Piston Gages. 1.2 The values stated in SI units are to be regarded as the standard. Other metric and inch-pound values are regarded as equivalent when required. 1.3 This practice is intended for the calibration of static force measuring instruments. It is not applicable for dynamic or high speed force calibrations, nor can the results of calibrations performed in accordance with this practice be assumed valid for...

  8. POLCAL - POLARIMETRIC RADAR CALIBRATION

    Science.gov (United States)

    Vanzyl, J.

    1994-01-01

    Calibration of polarimetric radar systems is a field of research in which great progress has been made over the last few years. POLCAL (Polarimetric Radar Calibration) is a software tool intended to assist in the calibration of Synthetic Aperture Radar (SAR) systems. In particular, POLCAL calibrates Stokes matrix format data produced as the standard product by the NASA/Jet Propulsion Laboratory (JPL) airborne imaging synthetic aperture radar (AIRSAR). POLCAL was designed to be used in conjunction with data collected by the NASA/JPL AIRSAR system. AIRSAR is a multifrequency (6 cm, 24 cm, and 68 cm wavelength), fully polarimetric SAR system which produces 12 x 12 km imagery at 10 m resolution. AIRSTAR was designed as a testbed for NASA's Spaceborne Imaging Radar program. While the images produced after 1991 are thought to be calibrated (phase calibrated, cross-talk removed, channel imbalance removed, and absolutely calibrated), POLCAL can and should still be used to check the accuracy of the calibration and to correct it if necessary. Version 4.0 of POLCAL is an upgrade of POLCAL version 2.0 released to AIRSAR investigators in June, 1990. New options in version 4.0 include automatic absolute calibration of 89/90 data, distributed target analysis, calibration of nearby scenes with calibration parameters from a scene with corner reflectors, altitude or roll angle corrections, and calibration of errors introduced by known topography. Many sources of error can lead to false conclusions about the nature of scatterers on the surface. Errors in the phase relationship between polarization channels result in incorrect synthesis of polarization states. Cross-talk, caused by imperfections in the radar antenna itself, can also lead to error. POLCAL reduces cross-talk and corrects phase calibration without the use of ground calibration equipment. Removing the antenna patterns during SAR processing also forms a very important part of the calibration of SAR data. Errors in the

  9. Characterization of the Sonoran desert as a radiometric calibration target for Earth observing sensors

    Science.gov (United States)

    Angal, Amit; Chander, Gyanesh; Xiong, Xiaoxiong; Choi, Tae-young; Wu, Aisheng

    2011-01-01

    To provide highly accurate quantitative measurements of the Earth's surface, a comprehensive calibration and validation of the satellite sensors is required. The NASA Moderate Resolution Imaging Spectroradiometer (MODIS) Characterization Support Team, in collaboration with United States Geological Survey, Earth Resources Observation and Science Center, has previously demonstrated the use of African desert sites to monitor the long-term calibration stability of Terra MODIS and Landsat 7 (L7) Enhanced Thematic Mapper plus (ETM+). The current study focuses on evaluating the suitability of the Sonoran Desert test site for post-launch long-term radiometric calibration as well as cross-calibration purposes. Due to the lack of historical and on-going in situ ground measurements, the Sonoran Desert is not usually used for absolute calibration. An in-depth evaluation (spatial, temporal, and spectral stability) of this site using well calibrated L7 ETM+ measurements and local climatology data has been performed. The Sonoran Desert site produced spatial variability of about 3 to 5% in the reflective solar regions, and the temporal variations of the site after correction for view-geometry impacts were generally around 3%. The results demonstrate that, barring the impacts due to occasional precipitation, the Sonoran Desert site can be effectively used for cross-calibration and long-term stability monitoring of satellite sensors, thus, providing a good test site in the western hemisphere.

  10. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  11. CLUSTER STAFF search coils magnetometer calibration - comparisons with FGM

    Science.gov (United States)

    Robert, P.; Cornilleau-Wehrlin, N.; Piberne, R.; de Conchy, Y.; Lacombe, C.; Bouzid, V.; Grison, B.; Alison, D.; Canu, P.

    2013-12-01

    The main part of Cluster Spatio Temporal Analysis of Field Fluctuations (STAFF) experiment consists of triaxial search coils allowing the measurements of the three magnetic components of the waves from 0.1 Hz up to 4 kHz. Two sets of data are produced, one by a module to filter and transmit the corresponding waveform up to either 10 or 180 Hz (STAFF-SC) and the second by an onboard Spectrum Analyser (STAFF-SA) to compute the elements of the spectral matrix for five components of the waves, 3 × B and 2 × E (from EFW experiment) in the frequency range 8 Hz to 4 kHz. In order to understand the way the output signal of the search coils are calibrated, the transfer functions of the different parts of the instrument are described as well as the way to transform telemetry data into physical units, across various coordinate systems from the spinning sensors to a fixed and known frame. The instrument sensitivity is discussed. Cross-calibration inside STAFF (SC and SA) is presented. Results of cross-calibration between the STAFF search coils and the Cluster Flux Gate Magnetometer (FGM) data are discussed. It is shown that these cross-calibrations lead to an agreement between both data sets at low frequency within a 2% error. By means of statistics done over 10 yr, it is shown that the functionalities and characteristics of both instruments have not changed during this period.

  12. CLUSTER-STAFF search coil magnetometer calibration - comparisons with FGM

    Science.gov (United States)

    Robert, P.; Cornilleau-Wehrlin, N.; Piberne, R.; de Conchy, Y.; Lacombe, C.; Bouzid, V.; Grison, B.; Alison, D.; Canu, P.

    2014-09-01

    The main part of the Cluster Spatio-Temporal Analysis of Field Fluctuations (STAFF) experiment consists of triaxial search coils allowing the measurements of the three magnetic components of the waves from 0.1 Hz up to 4 kHz. Two sets of data are produced, one by a module to filter and transmit the corresponding waveform up to either 10 or 180 Hz (STAFF-SC), and the second by the onboard Spectrum Analyser (STAFF-SA) to compute the elements of the spectral matrix for five components of the waves, 3 × B and 2 × E (from the EFW experiment), in the frequency range 8 Hz to 4 kHz. In order to understand the way the output signals of the search coils are calibrated, the transfer functions of the different parts of the instrument are described as well as the way to transform telemetry data into physical units across various coordinate systems from the spinning sensors to a fixed and known frame. The instrument sensitivity is discussed. Cross-calibration inside STAFF (SC and SA) is presented. Results of cross-calibration between the STAFF search coils and the Cluster Fluxgate Magnetometer (FGM) data are discussed. It is shown that these cross-calibrations lead to an agreement between both data sets at low frequency within a 2% error. By means of statistics done over 10 yr, it is shown that the functionalities and characteristics of both instruments have not changed during this period.

  13. Calibrated Properties Model

    International Nuclear Information System (INIS)

    Ahlers, C.; Liu, H.

    2000-01-01

    The purpose of this Analysis/Model Report (AMR) is to document the Calibrated Properties Model that provides calibrated parameter sets for unsaturated zone (UZ) flow and transport process models for the Yucca Mountain Site Characterization Project (YMP). This work was performed in accordance with the ''AMR Development Plan for U0035 Calibrated Properties Model REV00. These calibrated property sets include matrix and fracture parameters for the UZ Flow and Transport Model (UZ Model), drift seepage models, drift-scale and mountain-scale coupled-processes models, and Total System Performance Assessment (TSPA) models as well as Performance Assessment (PA) and other participating national laboratories and government agencies. These process models provide the necessary framework to test conceptual hypotheses of flow and transport at different scales and predict flow and transport behavior under a variety of climatic and thermal-loading conditions

  14. Calibration of Ga-68 activity for PET applications in Cuba

    International Nuclear Information System (INIS)

    García Rodríguez, Lourdes; Oropesa Verdecia, Pilar; Serra Águila, Rolando A.; Moreno León, Yecenia; Jénez Magaña, Yoel; Pérez LoretdeMola, Nayla; Bell Hechavarría, Ailec; Mas Ruiz, Javier; Cassette, Philippe

    2016-01-01

    A Ga-68 solution was used to calibrate the activity concentration using the double-triple coincidence ratio (TDCR) method of liquid scintillation for the first time in the country. The expanded uncertainty (k = 2) of the concentration of Ga-68 activity in the calibrated solution was equal to 2%. For measurements, the commercial liquid scintillation counter HIDEXTM was used. Samples were prepared by adding between 40 and 50 mg of the radioactive solution to 15 mL of ULTIMAGOLD ™ scintillating cocktail. For the estimation of Ga-68 counting efficiencies in the samples used for the calibration, a FORTRAN program developed by the National Institute of Metrology of France for the magnitudes of ionizing radiation, LNHB, was used. The validation of the method was carried out by the calibration of a standard solution of Na-22, also positronic emitter with similar disintegration scheme to Ga-68. The difference between the concentration of Na-22 activity measured using the TDCR method and the certified reference value traceable to the National Institute of Metrology of the United States (NIST) was 0.15%. With the solution of Ga-68 standardized by the TDCR method the calibration of the secondary standard activity meter, model CAPINTEC CRCTM 15R, was carried out for a geometry of 2R flask with 1mL of radioactive solution. Afterwards, this standard activity meter was calibrated for the measurement of Ga-68 in the geometries of interest in nuclear medicine: Flask 15R with 6 mL of radioactive solution, 2.5 mL syringe with 2 mL of radioactive solution and 5 mL syringe with 2 mL of radioactive solution. The results presented in this paper constitute the necessary metrological support for the introduction of new PET and PET / CT technologies into medical practice in Cuba.

  15. Calibration of high-dose radiation facilities (Handbook)

    International Nuclear Information System (INIS)

    Gupta, B.L.; Bhat, R.M.

    1986-01-01

    In India at present several high intensity radiation sources are used. There are 135 teletheraphy machines and 65 high intensity cobalt-60 sources in the form of gamma chambers (2.5 Ci) and PANBIT (50 Ci). Several food irradiation facilities and a medical sterilization plant ISOMED are also in operation. The application of these high intensity sources involve a wide variation of dose from 10 Gy to 100 kGy. Accurate and reproducible radiation dosimetry is essential in the use of these sources. This handbook is especially compiled for calibration of high-dose radiation facilities. The first few chapters discuss such topics as interaction of radiation with matter, radiation chemistry, radiation processing, commonly used high intensity radiation sources and their special features, radiation units and dosimetry principles. In the chapters which follow, chemical dosimeters are discussed in detail. This discussion covers Fricke dosimeter, FBX dosimeter, ceric sulphate dosimeter, free radical dosimetry, coloured indicators for irrdiation verification. A final chapter is devoted to practical hints to be followed in calibration work. (author)

  16. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR

    International Nuclear Information System (INIS)

    Acosta, Andy L. Romero; Lores, Stefan Gutierrez

    2013-01-01

    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba

  17. Calibration procedures of area monitors in terms of the Ambient Dose Equivalent H*(10), for gamma, x-ray radiation fields

    International Nuclear Information System (INIS)

    Dieguez Davila, L.E.

    1998-01-01

    In the present thesis procedures for calibrating portable survey meters in terms of the new ICRU quantities H*(10) ambient dose equivalent are discussed. Also the remendations of International Comission on Radiation Protection in their report ICRP 60 that inludes the operational magnitudes that the International Comission of Radiation Units proposed for calibrating area monitors

  18. Calibrated Properties Model

    International Nuclear Information System (INIS)

    Ahlers, C.F.; Liu, H.H.

    2001-01-01

    The purpose of this Analysis/Model Report (AMR) is to document the Calibrated Properties Model that provides calibrated parameter sets for unsaturated zone (UZ) flow and transport process models for the Yucca Mountain Site Characterization Project (YMP). This work was performed in accordance with the AMR Development Plan for U0035 Calibrated Properties Model REV00 (CRWMS M and O 1999c). These calibrated property sets include matrix and fracture parameters for the UZ Flow and Transport Model (UZ Model), drift seepage models, drift-scale and mountain-scale coupled-processes models, and Total System Performance Assessment (TSPA) models as well as Performance Assessment (PA) and other participating national laboratories and government agencies. These process models provide the necessary framework to test conceptual hypotheses of flow and transport at different scales and predict flow and transport behavior under a variety of climatic and thermal-loading conditions

  19. Radiometric Calibration of Osmi Imagery Using Solar Calibration

    Directory of Open Access Journals (Sweden)

    Dong-Han Lee

    2000-12-01

    Full Text Available OSMI (Ocean Scanning Multi-Spectral Imager raw image data (Level 0 were acquired and radiometrically corrected. We have applied two methods, using solar & dark calibration data from OSMI sensor and comparing with the SeaWiFS data, to the radiometric correction of OSMI raw image data. First, we could get the values of the gain and the offset for each pixel and each band from comparing the solar & dark calibration data with the solar input radiance values, calculated from the transmittance, BRDF (Bidirectional Reflectance Distribution Function and the solar incidence angle (¥â,¥è of OSMI sensor. Applying this calibration data to OSMI raw image data, we got the two odd results, the lower value of the radiometric corrected image data than the expected value, and the Venetian Blind Effect in the radiometric corrected image data. Second, we could get the reasonable results from comparing OSMI raw image data with the SeaWiFS data, and get a new problem of OSMI sensor.

  20. AFM lateral force calibration for an integrated probe using a calibration grating

    International Nuclear Information System (INIS)

    Wang, Huabin; Gee, Michelle L.

    2014-01-01

    Atomic force microscopy (AFM) friction measurements on hard and soft materials remain a challenge due to the difficulties associated with accurately calibrating the cantilever for lateral force measurement. One of the most widely accepted lateral force calibration methods is the wedge method. This method is often used in a simplified format but in so doing sacrifices accuracy. In the present work, we have further developed the wedge method to provide a lateral force calibration method for integrated AFM probes that is easy to use without compromising accuracy and reliability. Raw friction calibration data are collected from a single scan image by continuous ramping of the set point as the facets of a standard grating are scanned. These data are analysed to yield an accurate lateral force conversion/calibration factor that is not influenced by adhesion forces or load deviation. By demonstrating this new calibration method, we illustrate its reliability, speed and ease of execution. This method makes accessible reliable boundary lubrication studies on adhesive and heterogeneous surfaces that require spatial resolution of frictional forces. - Highlights: • We develop a simple and accurate method for lateral force calibration in AFM friction measurements. • We detail the basis of the method and illustrate how to use it and its reliability with example data. • Our method is easy, accurate and accounts for the affects of adhesion on friction measurements. • The method is applicable to integrated probes, as opposed to colloidal probes. • This allows accurate AFM friction measurements on spatially heterogeneous and adhesive surfaces

  1. A simple topography-driven, calibration-free runoff generation model

    Science.gov (United States)

    Gao, H.; Birkel, C.; Hrachowitz, M.; Tetzlaff, D.; Soulsby, C.; Savenije, H. H. G.

    2017-12-01

    Determining the amount of runoff generation from rainfall occupies a central place in rainfall-runoff modelling. Moreover, reading landscapes and developing calibration-free runoff generation models that adequately reflect land surface heterogeneities remains the focus of much hydrological research. In this study, we created a new method to estimate runoff generation - HAND-based Storage Capacity curve (HSC) which uses a topographic index (HAND, Height Above the Nearest Drainage) to identify hydrological similarity and partially the saturated areas of catchments. We then coupled the HSC model with the Mass Curve Technique (MCT) method to estimate root zone storage capacity (SuMax), and obtained the calibration-free runoff generation model HSC-MCT. Both the two models (HSC and HSC-MCT) allow us to estimate runoff generation and simultaneously visualize the spatial dynamic of saturated area. We tested the two models in the data-rich Bruntland Burn (BB) experimental catchment in Scotland with an unusual time series of the field-mapped saturation area extent. The models were subsequently tested in 323 MOPEX (Model Parameter Estimation Experiment) catchments in the United States. HBV and TOPMODEL were used as benchmarks. We found that the HSC performed better in reproducing the spatio-temporal pattern of the observed saturated areas in the BB catchment compared with TOPMODEL which is based on the topographic wetness index (TWI). The HSC also outperformed HBV and TOPMODEL in the MOPEX catchments for both calibration and validation. Despite having no calibrated parameters, the HSC-MCT model also performed comparably well with the calibrated HBV and TOPMODEL, highlighting the robustness of the HSC model to both describe the spatial distribution of the root zone storage capacity and the efficiency of the MCT method to estimate the SuMax. Moreover, the HSC-MCT model facilitated effective visualization of the saturated area, which has the potential to be used for broader

  2. Syringe calibration factors and volume correction factors for the NPL secondary standard radionuclide calibrator

    CERN Document Server

    Tyler, D K

    2002-01-01

    The activity assay of a radiopharmaceutical administration to a patient is normally achieved via the use of a radionuclide calibrator. Because of the different geometries and elemental compositions between plastic syringes and glass vials, the calibration factors for syringes may well be significantly different from those for the glass containers. The magnitude of these differences depends on the energies of the emitted photons. For some radionuclides variations have been observed of 70 %, it is therefore important to recalibrate for syringes or use syringe calibration factors. Calibration factors and volume correction factors have been derived for the NPL secondary standard radionuclide calibrator, for a variety of commonly used syringes and needles, for the most commonly used medical radionuclide.

  3. LiDAR-IMU Time Delay Calibration Based on Iterative Closest Point and Iterated Sigma Point Kalman Filter.

    Science.gov (United States)

    Liu, Wanli

    2017-03-08

    The time delay calibration between Light Detection and Ranging (LiDAR) and Inertial Measurement Units (IMUs) is an essential prerequisite for its applications. However, the correspondences between LiDAR and IMU measurements are usually unknown, and thus cannot be computed directly for the time delay calibration. In order to solve the problem of LiDAR-IMU time delay calibration, this paper presents a fusion method based on iterative closest point (ICP) and iterated sigma point Kalman filter (ISPKF), which combines the advantages of ICP and ISPKF. The ICP algorithm can precisely determine the unknown transformation between LiDAR-IMU; and the ISPKF algorithm can optimally estimate the time delay calibration parameters. First of all, the coordinate transformation from the LiDAR frame to the IMU frame is realized. Second, the measurement model and time delay error model of LiDAR and IMU are established. Third, the methodology of the ICP and ISPKF procedure is presented for LiDAR-IMU time delay calibration. Experimental results are presented that validate the proposed method and demonstrate the time delay error can be accurately calibrated.

  4. RF impedance measurement calibration

    International Nuclear Information System (INIS)

    Matthews, P.J.; Song, J.J.

    1993-01-01

    The intent of this note is not to explain all of the available calibration methods in detail. Instead, we will focus on the calibration methods of interest for RF impedance coupling measurements and attempt to explain: (1). The standards and measurements necessary for the various calibration techniques. (2). The advantages and disadvantages of each technique. (3). The mathematical manipulations that need to be applied to the measured standards and devices. (4). An outline of the steps needed for writing a calibration routine that operated from a remote computer. For further details of the various techniques presented in this note, the reader should consult the references

  5. Robot-assisted automatic ultrasound calibration.

    Science.gov (United States)

    Aalamifar, Fereshteh; Cheng, Alexis; Kim, Younsu; Hu, Xiao; Zhang, Haichong K; Guo, Xiaoyu; Boctor, Emad M

    2016-10-01

    Ultrasound (US) calibration is the process of determining the unknown transformation from a coordinate frame such as the robot's tooltip to the US image frame and is a necessary task for any robotic or tracked US system. US calibration requires submillimeter-range accuracy for most applications, but it is a time-consuming and repetitive task. We provide a new framework for automatic US calibration with robot assistance and without the need for temporal calibration. US calibration based on active echo (AE) phantom was previously proposed, and its superiority over conventional cross-wire phantom-based calibration was shown. In this work, we use AE to guide the robotic arm motion through the process of data collection; we combine the capability of the AE point to localize itself in the frame of the US image with the automatic motion of the robotic arm to provide a framework for calibrating the arm to the US image automatically. We demonstrated the efficacy of the automated method compared to the manual method through experiments. To highlight the necessity of frequent ultrasound calibration, it is demonstrated that the calibration precision changed from 1.67 to 3.20 mm if the data collection is not repeated after a dismounting/mounting of the probe holder. In a large data set experiment, similar reconstruction precision of automatic and manual data collection was observed, while the time was reduced by 58 %. In addition, we compared ten automatic calibrations with ten manual ones, each performed in 15 min, and showed that all the automatic ones could converge in the case of setting the initial matrix as identity, while this was not achieved by manual data sets. Given the same initial matrix, the repeatability of the automatic was [0.46, 0.34, 0.80, 0.47] versus [0.42, 0.51, 0.98, 1.15] mm in the manual case for the US image four corners. The submillimeter accuracy requirement of US calibration makes frequent data collections unavoidable. We proposed an automated

  6. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  7. Calibration of the fluxgate CSC vector magnetometers

    DEFF Research Database (Denmark)

    Merayo, José M.G.; Risbo, Torben; Primdahl, Fritz

    1995-01-01

    sensor giving an error corresponding to 5 to 10 nT in full scale field. It rotates as well the CSC sensor axes. In principle this is due to the presence of soft magnetic material. No remanent magnetization has been observed. Hermann Luehr has ratified this fact and recommended a final calibration...... thermal run of the flight unit. The residuals show some outliers due to the mechanical perturbation while removing the dry ice. Due to the speed of temperature change the thermodynamic equilibrium is not well achieved in the cooling phase, and that gives a little bit dispersion of the parameters....

  8. A Study on Relative Radiometric Calibration without Calibration Field for YG-25

    Directory of Open Access Journals (Sweden)

    ZHANG Guo

    2017-08-01

    Full Text Available YG-25 is the first agility optical remote sensing satellite of China to acquire the sub-meter imagery of the earth. The side slither calibration technique is an on-orbit maneuver that has been used to flat-field image data acquired over the uniform calibration field. However, imaging to the single uniform calibration field cannot afford to calibrate the full dynamic response range of the sensor and reduces the efficiency. The paper proposes a new relative radiometric calibration method that a 90-degree yaw maneuver is performed over any non-uniform features of the Earth for YG-25. Meanwhile, we use an enhanced side slither image horizontal correction method based on line segment detector(LSDalgorithm to solve the side slither image over-shifted problem.The shifted results are compared with other horizontal correction method. The histogram match algorithm is used to calculate the relative gains of all detectors. The correctness and validity of the proposed method are validated by using the YG-25 on-board side slither data. The results prove that the mean streaking metrics of relative correction images of YG-25 is better 0.07%, the noticeable striping artifact and residual noise are removed, the calibration accuracy of side slither technique based on non-uniform features is superior to life image statistics of sensor's life span.

  9. Pulse-based internal calibration of polarimetric SAR

    DEFF Research Database (Denmark)

    Dall, Jørgen; Skou, Niels; Christensen, Erik Lintz

    1994-01-01

    Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops and devel......Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops...

  10. Site Calibration report

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Vesth, Allan

    The report describes site calibration measurements carried out on a site in Denmark. The measurements are carried out in accordance to Ref. [1]. The site calibration is carried out before a power performance measurement on a given turbine to clarify the influence from the terrain on the ratio...

  11. LANL MTI calibration team experience

    Science.gov (United States)

    Bender, Steven C.; Atkins, William H.; Clodius, William B.; Little, Cynthia K.; Christensen, R. Wynn

    2004-01-01

    The Multispectral Thermal Imager (MTI) was designed as an imaging radiometer with absolute calibration requirements established by Department of Energy (DOE) mission goals. Particular emphasis was given to water surface temperature retrieval using two mid wave and three long wave infrared spectral bands, the fundamental requirement was a surface temperature determination of 1K at the 68% confidence level. For the ten solar reflective bands a one-sigma radiometric performance goal of 3% was established. In order to address these technical challenges a calibration facility was constructed containing newly designed sources that were calibrated at NIST. Additionally, the design of the payload and its onboard calibration system supported post launch maintenance and update of the ground calibration. The on-orbit calibration philosophy also included vicarious techniques using ocean buoys, playas and other instrumented sites; these became increasingly important subsequent to an electrical failure which disabled the onboard calibration system. This paper offers various relevant lessons learned in the eight-year process of reducing to practice the calibration capability required by the scientific mission. The discussion presented will include observations pertinent to operational and procedural issues as well as hardware experiences; the validity of some of the initial assumptions will also be explored.

  12. Vibration transducer calibration techniques

    Science.gov (United States)

    Brinkley, D. J.

    1980-09-01

    Techniques for the calibration of vibration transducers used in the Aeronautical Quality Assurance Directorate of the British Ministry of Defence are presented. Following a review of the types of measurements necessary in the calibration of vibration transducers, the performance requirements of vibration transducers, which can be used to measure acceleration, velocity or vibration amplitude, are discussed, with particular attention given to the piezoelectric accelerometer. Techniques for the accurate measurement of sinusoidal vibration amplitude in reference-grade transducers are then considered, including the use of a position sensitive photocell and the use of a Michelson laser interferometer. Means of comparing the output of working-grade accelerometers with that of previously calibrated reference-grade devices are then outlined, with attention given to a method employing a capacitance bridge technique and a method to be used at temperatures between -50 and 200 C. Automatic calibration procedures developed to speed up the calibration process are outlined, and future possible extensions of system software are indicated.

  13. Energy calibration issues in nuclear resonant vibrational spectroscopy: observing small spectral shifts and making fast calibrations.

    Science.gov (United States)

    Wang, Hongxin; Yoda, Yoshitaka; Dong, Weibing; Huang, Songping D

    2013-09-01

    The conventional energy calibration for nuclear resonant vibrational spectroscopy (NRVS) is usually long. Meanwhile, taking NRVS samples out of the cryostat increases the chance of sample damage, which makes it impossible to carry out an energy calibration during one NRVS measurement. In this study, by manipulating the 14.4 keV beam through the main measurement chamber without moving out the NRVS sample, two alternative calibration procedures have been proposed and established: (i) an in situ calibration procedure, which measures the main NRVS sample at stage A and the calibration sample at stage B simultaneously, and calibrates the energies for observing extremely small spectral shifts; for example, the 0.3 meV energy shift between the 100%-(57)Fe-enriched [Fe4S4Cl4](=) and 10%-(57)Fe and 90%-(54)Fe labeled [Fe4S4Cl4](=) has been well resolved; (ii) a quick-switching energy calibration procedure, which reduces each calibration time from 3-4 h to about 30 min. Although the quick-switching calibration is not in situ, it is suitable for normal NRVS measurements.

  14. On chromatic and geometrical calibration

    DEFF Research Database (Denmark)

    Folm-Hansen, Jørgen

    1999-01-01

    The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... the correct interpolation method is described. For the chromatic issues of calibration we present the acquisition of colour and multi-spectral images, the chromatic aberrations and the various lens/camera based non-uniformities of the illumination of the image plane. It is described how the monochromatic...... to design calibration targets for both geometrical and chromatic calibration are described. We present some possible systematical errors on the detection of the objects in the calibration targets, if viewed in a non orthogonal angle, if the intensities are uneven or if the image blurring is uneven. Finally...

  15. Calibration of triaxial fluxgate gradiometer

    International Nuclear Information System (INIS)

    Vcelak, Jan

    2006-01-01

    The description of simple and fast calibration procedures used for double-probe triaxial fluxgate gradiometer is provided in this paper. The calibration procedure consists of three basic steps. In the first step both probes are calibrated independently in order to reach constant total field reading in every position. Both probes are numerically aligned in the second step in order that the gradient reading is zero in homogenous magnetic field. The third step consists of periodic drift calibration during measurement. The results and detailed description of each calibration step are presented and discussed in the paper. The gradiometer is finally verified during the detection of the metal object in the measuring grid

  16. A round-robin gamma stereotactic radiosurgery dosimetry interinstitution comparison of calibration protocols

    Energy Technology Data Exchange (ETDEWEB)

    Drzymala, R. E., E-mail: drzymala@wustl.edu [Department of Radiation Oncology, Washington University, St. Louis, Missouri 63110 (United States); Alvarez, P. E. [Imaging and Radiation Oncology Core Houston, UT MD Anderson Cancer Center, Houston, Texas 77030 (United States); Bednarz, G. [Radiation Oncology Department, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania 15232 (United States); Bourland, J. D. [Department of Radiation Oncology, Wake Forest University, Winston-Salem, North Carolina 27157 (United States); DeWerd, L. A. [Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53705 (United States); Ma, L. [Department of Radiation Oncology, University California San Francisco, San Francisco, California 94143 (United States); Meltsner, S. G. [Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina 27710 (United States); Neyman, G. [Department of Radiation Oncology, The Cleveland Clinic Foundation, Cleveland, Ohio 44195 (United States); Novotny, J. [Medical Physics Department, Hospital Na Homolce, Prague 15030 (Czech Republic); Petti, P. L. [Gamma Knife Center, Washington Hospital Healthcare System, Fremont, California 94538 (United States); Rivard, M. J. [Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111 (United States); Shiu, A. S. [Department of Radiation Oncology, University of Southern California, Los Angeles, California 90033 (United States); Goetsch, S. J. [San Diego Medical Physics, Inc., La Jolla, California 92037 (United States)

    2015-11-15

    Purpose: Absorbed dose calibration for gamma stereotactic radiosurgery is challenging due to the unique geometric conditions, dosimetry characteristics, and nonstandard field size of these devices. Members of the American Association of Physicists in Medicine (AAPM) Task Group 178 on Gamma Stereotactic Radiosurgery Dosimetry and Quality Assurance have participated in a round-robin exchange of calibrated measurement instrumentation and phantoms exploring two approved and two proposed calibration protocols or formalisms on ten gamma radiosurgery units. The objectives of this study were to benchmark and compare new formalisms to existing calibration methods, while maintaining traceability to U.S. primary dosimetry calibration laboratory standards. Methods: Nine institutions made measurements using ten gamma stereotactic radiosurgery units in three different 160 mm diameter spherical phantoms [acrylonitrile butadiene styrene (ABS) plastic, Solid Water, and liquid water] and in air using a positioning jig. Two calibrated miniature ionization chambers and one calibrated electrometer were circulated for all measurements. Reference dose-rates at the phantom center were determined using the well-established AAPM TG-21 or TG-51 dose calibration protocols and using two proposed dose calibration protocols/formalisms: an in-air protocol and a formalism proposed by the International Atomic Energy Agency (IAEA) working group for small and nonstandard radiation fields. Each institution’s results were normalized to the dose-rate determined at that institution using the TG-21 protocol in the ABS phantom. Results: Percentages of dose-rates within 1.5% of the reference dose-rate (TG-21 + ABS phantom) for the eight chamber-protocol-phantom combinations were the following: 88% for TG-21, 70% for TG-51, 93% for the new IAEA nonstandard-field formalism, and 65% for the new in-air protocol. Averages and standard deviations for dose-rates over all measurements relative to the TG-21 + ABS

  17. Calibrating Legal Judgments

    Directory of Open Access Journals (Sweden)

    Frederick Schauer

    2017-09-01

    Full Text Available Objective to study the notion and essence of legal judgments calibration the possibilities of using it in the lawenforcement activity to explore the expenses and advantages of using it. Methods dialectic approach to the cognition of social phenomena which enables to analyze them in historical development and functioning in the context of the integrity of objective and subjective factors it determined the choice of the following research methods formallegal comparative legal sociological methods of cognitive psychology and philosophy. Results In ordinary life people who assess other peoplersaquos judgments typically take into account the other judgments of those they are assessing in order to calibrate the judgment presently being assessed. The restaurant and hotel rating website TripAdvisor is exemplary because it facilitates calibration by providing access to a raterrsaquos previous ratings. Such information allows a user to see whether a particular rating comes from a rater who is enthusiastic about every place she patronizes or instead from someone who is incessantly hard to please. And even when less systematized as in assessing a letter of recommendation or college transcript calibration by recourse to the decisional history of those whose judgments are being assessed is ubiquitous. Yet despite the ubiquity and utility of such calibration the legal system seems perversely to reject it. Appellate courts do not openly adjust their standard of review based on the previous judgments of the judge whose decision they are reviewing nor do judges in reviewing legislative or administrative decisions magistrates in evaluating search warrant representations or jurors in assessing witness perception. In most legal domains calibration by reference to the prior decisions of the reviewee is invisible either because it does not exist or because reviewing bodies are unwilling to admit using what they in fact know and employ. Scientific novelty for the first

  18. Upper Limb Kinematics Using Inertial and Magnetic Sensors: Comparison of Sensor-to-Segment Calibrations

    Directory of Open Access Journals (Sweden)

    Brice Bouvier

    2015-07-01

    Full Text Available Magneto-Inertial Measurement Unit sensors (MIMU display high potential for the quantitative evaluation of upper limb kinematics, as they allow monitoring ambulatory measurements. The sensor-to-segment calibration step, consisting of establishing the relation between MIMU sensors and human segments, plays an important role in the global accuracy of joint angles. The aim of this study was to compare sensor-to-segment calibrations for the MIMU-based estimation of wrist, elbow, and shoulder joint angles, by examining trueness (“close to the reference” and precision (reproducibility validity criteria. Ten subjects performed five sessions with three different operators. Three classes of calibrations were studied: segment axes equal to technical MIMU axes (TECH, segment axes generated during a static pose (STATIC, and those generated during functional movements (FUNCT. The calibrations were compared during the maximal uniaxial movements of each joint, plus an extra multi-joint movement. Generally, joint angles presented good trueness and very good precision in the range 5°–10°. Only small discrepancy between calibrations was highlighted, with the exception of a few cases. The very good overall accuracy (trueness and precision of MIMU-based joint angle data seems to be more dependent on the level of rigor of the experimental procedure (operator training than on the choice of calibration itself.

  19. United States Army Aeromedical Support to African Fliers, 1941 - 1949: The Tuskegee Flight Suregons

    National Research Council Canada - National Science Library

    Jones, David R; Gross, Leroy P; Marchbanks-Robinson, Roslyn

    2007-01-01

    .... Segregation policies in early 1941 required a few African American flight surgeons to support black student pilots at Tuskegee, Alabama from cadet training in preflight ground school until graduation with pilot wings...

  20. Optimal Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard; Kroon, I. B.; Faber, Michael Havbro

    1994-01-01

    Calibration of partial safety factors is considered in general, including classes of structures where no code exists beforehand. The partial safety factors are determined such that the difference between the reliability for the different structures in the class considered and a target reliability...... level is minimized. Code calibration on a decision theoretical basis is also considered and it is shown how target reliability indices can be calibrated. Results from code calibration for rubble mound breakwater designs are shown....

  1. Observation models in radiocarbon calibration

    International Nuclear Information System (INIS)

    Jones, M.D.; Nicholls, G.K.

    2001-01-01

    The observation model underlying any calibration process dictates the precise mathematical details of the calibration calculations. Accordingly it is important that an appropriate observation model is used. Here this is illustrated with reference to the use of reservoir offsets where the standard calibration approach is based on a different model to that which the practitioners clearly believe is being applied. This sort of error can give rise to significantly erroneous calibration results. (author). 12 refs., 1 fig

  2. A calibration method for proposed XRF measurements of arsenic and selenium in nail clippings

    International Nuclear Information System (INIS)

    Gherase, Mihai R; Fleming, David E B

    2011-01-01

    A calibration method for proposed x-ray fluorescence (XRF) measurements of arsenic and selenium in nail clippings is demonstrated. Phantom nail clippings were produced from a whole nail phantom (0.7 mm thickness, 25 x 25 mm 2 area) and contained equal concentrations of arsenic and selenium ranging from 0 to 20 μg g -1 in increments of 5 μg g -1 . The phantom nail clippings were then grouped in samples of five different masses: 20, 40, 60, 80 and 100 mg for each concentration. Experimental x-ray spectra were acquired for each of the sample masses using a portable x-ray tube and a detector unit. Calibration lines (XRF signal in a number of counts versus stoichiometric elemental concentration) were produced for each of the two elements. A semi-empirical relationship between the mass of the nail phantoms (m) and the slope of the calibration line (s) was determined separately for arsenic and selenium. Using this calibration method, one can estimate elemental concentrations and their uncertainties from the XRF spectra of human nail clippings. (note)

  3. Technical Note: Procedure for the calibration and validation of kilo-voltage cone-beam CT models

    Energy Technology Data Exchange (ETDEWEB)

    Vilches-Freixas, Gloria; Létang, Jean Michel; Rit, Simon, E-mail: simon.rit@creatis.insa-lyon.fr [Université de Lyon, CREATIS, CNRS UMR5220, Inserm U1206, INSA-Lyon, Université Lyon 1, Centre Léon Bérard, Lyon 69373 Cedex 08 (France); Brousmiche, Sébastien [Ion Beam Application, Louvain-la-Neuve 1348 (Belgium); Romero, Edward; Vila Oliva, Marc [Université de Lyon, CREATIS, CNRS UMR5220, Inserm U1206, INSA-Lyon, Université Lyon 1, Centre Léon Bérard, Lyon 69373 Cedex 08, France and Ion Beam Application, Louvain-la-Neuve 1348 (Belgium); Kellner, Daniel; Deutschmann, Heinz; Keuschnigg, Peter; Steininger, Philipp [Institute for Research and Development on Advanced Radiation Technologies, Paracelsus Medical University, Salzburg 5020 (Austria)

    2016-09-15

    Purpose: The aim of this work is to propose a general and simple procedure for the calibration and validation of kilo-voltage cone-beam CT (kV CBCT) models against experimental data. Methods: The calibration and validation of the CT model is a two-step procedure: the source model then the detector model. The source is described by the direction dependent photon energy spectrum at each voltage while the detector is described by the pixel intensity value as a function of the direction and the energy of incident photons. The measurements for the source consist of a series of dose measurements in air performed at each voltage with varying filter thicknesses and materials in front of the x-ray tube. The measurements for the detector are acquisitions of projection images using the same filters and several tube voltages. The proposed procedure has been applied to calibrate and assess the accuracy of simple models of the source and the detector of three commercial kV CBCT units. If the CBCT system models had been calibrated differently, the current procedure would have been exclusively used to validate the models. Several high-purity attenuation filters of aluminum, copper, and silver combined with a dosimeter which is sensitive to the range of voltages of interest were used. A sensitivity analysis of the model has also been conducted for each parameter of the source and the detector models. Results: Average deviations between experimental and theoretical dose values are below 1.5% after calibration for the three x-ray sources. The predicted energy deposited in the detector agrees with experimental data within 4% for all imaging systems. Conclusions: The authors developed and applied an experimental procedure to calibrate and validate any model of the source and the detector of a CBCT unit. The present protocol has been successfully applied to three x-ray imaging systems. The minimum requirements in terms of material and equipment would make its implementation suitable in

  4. Landsat-8 Sensor Characterization and Calibration

    Directory of Open Access Journals (Sweden)

    Brian Markham

    2015-02-01

    Full Text Available Landsat-8 was launched on 11 February 2013 with two new Earth Imaging sensors to provide a continued data record with the previous Landsats. For Landsat-8, pushbroom technology was adopted, and the reflective bands and thermal bands were split into two instruments. The Operational Land Imager (OLI is the reflective band sensor and the Thermal Infrared Sensor (TIRS, the thermal. In addition to these fundamental changes, bands were added, spectral bandpasses were refined, dynamic range and data quantization were improved, and numerous other enhancements were implemented. As in previous Landsat missions, the National Aeronautics and Space Administration (NASA and United States Geological Survey (USGS cooperated in the development, launch and operation of the Landsat-8 mission. One key aspect of this cooperation was in the characterization and calibration of the instruments and their data. This Special Issue documents the efforts of the joint USGS and NASA calibration team and affiliates to characterize the new sensors and their data for the benefit of the scientific and application users of the Landsat archive. A key scientific use of Landsat data is to assess changes in the land-use and land cover of the Earth’s surface over the now 43-year record. [...

  5. Calibration simulation. A calibration Monte-Carlo program for the OPAL jet chamber

    International Nuclear Information System (INIS)

    Biebel, O.

    1989-12-01

    A calibration Monte Carlo program has been developed as a tool to investigate the interdependence of track reconstruction and calibration constants. Three categories of calibration effects have been considered: The precise knowledge of sense wire positions, necessary to reconstruct the particle trajectories in the jet chamber. Included are the staggering and the sag of the sense wires as well as tilts and rotations of their support structures. The various contributions to the measured drift time, with special emphasis on the aberration due to the track angle and the presence of a transverse magnetic field. A very precise knowledge of the drift velocity and the Lorentz angle of the drift paths with respect to the drift field is also required. The effects degrading particle identification via energy loss dE/dx. Impurities of the gas mixture and saturation effects depending on the track angle as well as the influence of the pulse shaping-electronics have been studied. These effects have been parametrised with coefficients corresponding to the calibration constants required for track reconstruction. Excellent agreement with the input data has been achieved when determining calibration constants from Monte Carlo data generated with these parametrisations. (orig.) [de

  6. New Submersed Chamber for Calibration of Relative Humidity Instruments at HMI/FSB-LPM

    Science.gov (United States)

    Sestan, D.; Zvizdic, D.; Sariri, K.

    2018-02-01

    This paper gives a detailed description of a new chamber designed for calibration of relative humidity (RH) instruments at Laboratory for Process Measurement (HMI/FSB-LPM). To the present time, the calibrations of RH instruments at the HMI/FSB-LPM were done by comparison method using a climatic chamber of large volume and calibrated dew point hygrometer with an additional thermometer. Since 2010, HMI/FSB-LPM in cooperation with Centre for Metrology and Accreditation in Finland (MIKES) developed the two primary dew point generators which cover the dew point temperature range between - 70 {°}C and 60 {°}C. In order to utilize these facilities for calibrations of the RH instruments, the new chamber was designed, manufactured and installed in the existing system, aiming to extend its range and reduce the related calibration uncertainties. The chamber construction allows its use in a thermostatic bath of larger volume as well as in the climatic chambers. In the scope of this paper, performances of the new chamber were tested while it was submersed in a thermostated bath. The chamber can simultaneously accommodate up to three RH sensors. In order to keep the design of the chamber simple, only cylindrical RH sensors detachable from display units can be calibrated. Possible optimizations are also discussed, and improvements in the design proposed. By using the new chamber, HMI/FSB-LPM reduced the expanded calibration uncertainties (level of confidence 95 %, coverage factor k=2) from 0.6 %rh to 0.25 %rh at 30 %rh (23 {°}C), and from 0.8 %rh to 0.53 %rh at 70 %rh (23 {°}C).

  7. Portable compact multifunction IR calibrator

    International Nuclear Information System (INIS)

    Wyatt, C.L.; Jacobsen, L.; Steed, A.

    1988-01-01

    A compact portable multifunction calibrator designed for future sensor systems is described which enables a linearity calibration for all detectors simultaneously using a near small-area source, a high-resolution mapping of the focal plane with 10 microrad setability and with a blur of less than 100 microrad, system spectral response calibration (radiometer) using a Michelson interferometer source, relative spectral response (spectrometer) using high-temperature external commercial blackbody simulators, and an absolute calibration using an internal low-temperature extended-area source. 5 references

  8. Calibration of intense 60Co gamma ray sources at the Savannah River Plant

    International Nuclear Information System (INIS)

    Bibler, N.E.

    1976-05-01

    Three different dosimeters were used to calibrate Savannah River Plant 60 Co sources having intensities greater than 10 7 rads/hr. These dosimeters are (a) ceric sulfate dissolved in 0.4M H 2 SO 4 , (b) oxalic acid dissolved in water, and (c) a commercially available nylon film containing a radiochromic dye. Response per unit dose to these dosimeters is independent of radiation intensity at 10 4 to 10 11 rads/hr. The dosimeters were calibrated at 6.0 x 10 5 rads/hr with a 60 Co source whose intensity was determined with the standard Fricke dosimeter. For the sources at 10 7 rads/hr or greater, intensities were calculated from slopes of linear plots of dosimeter response versus irradiation time. Individual dose rates varied from 1.0 x 10 7 to 4.6 x 10 7 rads/hr. Each source was calibrated with at least two different dosimeters. Relative standard deviations varied from 2 to 9 percent. A conservative estimate of the uncertainty in the accuracy of these dosimeters is 10 percent. Of the three dosimeters, the nylon film is easiest to use and is therefore recommended for future calibrations

  9. Development of portable flow calibrator

    International Nuclear Information System (INIS)

    Akiyama, Kiyomitsu; Iijima, Nobuo

    1995-01-01

    In the nuclear facilities, air sniffer system is often utilized to evaluate atmospheric concentration of radioactivity in the working environment. The system collects airborne dust on the filter during some sampling period. In this method, total air flow during the sampling period is an important parameter to evaluate the radioactivity concentration correctly. Therefore, calibration for the flow meter of air sniffer system must be done periodically according to Japan Industry Standards (JIS). As we have had to available device to calibrate the flow meter in the working area, we had to remove the flow meters from the installed place and carry them to another place where calibration can be made. This work required a great deal of labor. Now we have developed a portable flow calibrator for air sniffer system which enables us to make in-site calibration of the flow meter in the working area more easily. This report describes the outline of portable flow calibrator and it's experimental results. (author)

  10. Standardization of Laser Methods and Techniques for Vibration Measurements and Calibrations

    International Nuclear Information System (INIS)

    Martens, Hans-Juergen von

    2010-01-01

    The realization and dissemination of the SI units of motion quantities (vibration and shock) have been based on laser interferometer methods specified in international documentary standards. New and refined laser methods and techniques developed by national metrology institutes and by leading manufacturers in the past two decades have been swiftly specified as standard methods for inclusion into in the series ISO 16063 of international documentary standards. A survey of ISO Standards for the calibration of vibration and shock transducers demonstrates the extended ranges and improved accuracy (measurement uncertainty) of laser methods and techniques for vibration and shock measurements and calibrations. The first standard for the calibration of laser vibrometers by laser interferometry or by a reference accelerometer calibrated by laser interferometry (ISO 16063-41) is on the stage of a Draft International Standard (DIS) and may be issued by the end of 2010. The standard methods with refined techniques proved to achieve wider measurement ranges and smaller measurement uncertainties than that specified in the ISO Standards. The applicability of different standardized interferometer methods to vibrations at high frequencies was recently demonstrated up to 347 kHz (acceleration amplitudes up to 350 km/s 2 ). The relative deviations between the amplitude measurement results of the different interferometer methods that were applied simultaneously, differed by less than 1% in all cases.

  11. An Expectation-Maximization Method for Calibrating Synchronous Machine Models

    Energy Technology Data Exchange (ETDEWEB)

    Meng, Da; Zhou, Ning; Lu, Shuai; Lin, Guang

    2013-07-21

    The accuracy of a power system dynamic model is essential to its secure and efficient operation. Lower confidence in model accuracy usually leads to conservative operation and lowers asset usage. To improve model accuracy, this paper proposes an expectation-maximization (EM) method to calibrate the synchronous machine model using phasor measurement unit (PMU) data. First, an extended Kalman filter (EKF) is applied to estimate the dynamic states using measurement data. Then, the parameters are calculated based on the estimated states using maximum likelihood estimation (MLE) method. The EM method iterates over the preceding two steps to improve estimation accuracy. The proposed EM method’s performance is evaluated using a single-machine infinite bus system and compared with a method where both state and parameters are estimated using an EKF method. Sensitivity studies of the parameter calibration using EM method are also presented to show the robustness of the proposed method for different levels of measurement noise and initial parameter uncertainty.

  12. Fuel conditioning facility electrorefiner volume calibration

    International Nuclear Information System (INIS)

    Bucher, R.G.; Orechwa, Y.

    1995-01-01

    In one of the electrometallurgical process steps of the Fuel Conditioning Facility (FCF), die in-process nuclear material is dissolved in the electrorefiner tank in an upper layer of a mixture of liquid LiCl-KCl salt and a lower layer of liquid cadmium. The electrorefiner tank, as most process tanks, is not a smooth right-circular cylinder for which a single linear volume calibration curve could be fitted over the whole height of the tank. Rather, the tank contains many internal components, which cause systematic deviations from a single linear function. The nominal operating temperature of the electrorefiner is 500 degrees C although the salt and cadmium are introduced at 410 degrees C. The operating materials and temperatures preclude multiple calibration runs at operating conditions. In order to maximize the calibration information, multiple calibration runs were performed with water at room temperature. These data allow identification of calibration segments, and preliminary estimation of the calibration function and calibration uncertainties. The final calibration function is based on a combination of data from die water calibrations and the measurements made during the filling of the electrorefiner with salt and cadmium for operation

  13. NIST high-dose calibration services

    International Nuclear Information System (INIS)

    Humphreys, J.C.

    1989-01-01

    There is a need for the standardization of high-dose measurements used in the radiation-processing industry in order to provide assured traceability to national standards. NIST provides dosimetry calibration services to this industry. One of these services involves administration of known absorbed doses of gamma rays to customer-supplied dosimeters. The dosimeters are packaged to provide electron equilibrium conditions and are irradiated in a standard 60 Co calibration facility; this provides a calibration of that batch of dosimeters. Another service consists of supplying to a customer calibrated transfer dosimeters for irradiation with the customer's radiation source. The irradiated transfer dosimeters are then returned to NIST for analysis; the results are reported to the customer, providing a calibration of the dose rate of the customer's source. (orig.)

  14. TIME CALIBRATED OSCILLOSCOPE SWEEP CIRCUIT

    Science.gov (United States)

    Smith, V.L.; Carstensen, H.K.

    1959-11-24

    An improved time calibrated sweep circuit is presented, which extends the range of usefulness of conventional oscilloscopes as utilized for time calibrated display applications in accordance with U. S. Patent No. 2,832,002. Principal novelty resides in the provision of a pair of separate signal paths, each of which is phase and amplitude adjustable, to connect a high-frequency calibration oscillator to the output of a sawtooth generator also connected to the respective horizontal deflection plates of an oscilloscope cathode ray tube. The amplitude and phase of the calibration oscillator signals in the two signal paths are adjusted to balance out feedthrough currents capacitively coupled at high frequencies of the calibration oscillator from each horizontal deflection plate to the vertical plates of the cathode ray tube.

  15. Lidar to lidar calibration

    DEFF Research Database (Denmark)

    Fernandez Garcia, Sergio; Villanueva, Héctor

    This report presents the result of the lidar to lidar calibration performed for ground-based lidar. Calibration is here understood as the establishment of a relation between the reference lidar wind speed measurements with measurement uncertainties provided by measurement standard and corresponding...... lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from the reference lidar measurements are given for information only....

  16. Self-Calibration of CMB Polarimeters

    Science.gov (United States)

    Keating, Brian

    2013-01-01

    Precision measurements of the polarization of the cosmic microwave background (CMB) radiation, especially experiments seeking to detect the odd-parity "B-modes", have far-reaching implications for cosmology. To detect the B-modes generated during inflation the flux response and polarization angle of these experiments must be calibrated to exquisite precision. While suitable flux calibration sources abound, polarization angle calibrators are deficient in many respects. Man-made polarized sources are often not located in the antenna's far-field, have spectral properties that are radically different from the CMB's, are cumbersome to implement and may be inherently unstable over the (long) duration these searches require to detect the faint signature of the inflationary epoch. Astrophysical sources suffer from time, frequency and spatial variability, are not visible from all CMB observatories, and none are understood with sufficient accuracy to calibrate future CMB polarimeters seeking to probe inflationary energy scales of ~1000 TeV. CMB TB and EB modes, expected to identically vanish in the standard cosmological model, can be used to calibrate CMB polarimeters. By enforcing the observed EB and TB power spectra to be consistent with zero, CMB polarimeters can be calibrated to levels not possible with man-made or astrophysical sources. All of this can be accomplished without any loss of observing time using a calibration source which is spectrally identical to the CMB B-modes. The calibration procedure outlined here can be used for any CMB polarimeter.

  17. The dissemination in France of the units used in ionizing radiation metrology

    International Nuclear Information System (INIS)

    Guiho, J.P.; Simoen, J.P.

    1978-01-01

    After reviewing the system of metrology in France the authors describe the working of the ionizing radiation calibration chain. Emphasis is laid on the procedures used for the transfer of the units of exposure and absorbed dose. Such transfers are carried out either by direct comparison with a standard kept at the calibration centre, or by special procedures involving the use of transfer dose meters or of radioactive sources calibrated and supplied by the primary laboratory or the calibration centre. An analysis of the steps and of the accumulation of errors is presented for each dosimetric quantity considered. The authors make a preliminary assessment of the operation of the French ionizing radiation calibration chain. (author)

  18. High Gain Antenna Calibration on Three Spacecraft

    Science.gov (United States)

    Hashmall, Joseph A.

    2011-01-01

    This paper describes the alignment calibration of spacecraft High Gain Antennas (HGAs) for three missions. For two of the missions (the Lunar Reconnaissance Orbiter and the Solar Dynamics Observatory) the calibration was performed on orbit. For the third mission (the Global Precipitation Measurement core satellite) ground simulation of the calibration was performed in a calibration feasibility study. These three satellites provide a range of calibration situations-Lunar orbit transmitting to a ground antenna for LRO, geosynchronous orbit transmitting to a ground antenna fer SDO, and low Earth orbit transmitting to TDRS satellites for GPM The calibration results depend strongly on the quality and quantity of calibration data. With insufficient data the calibration Junction may give erroneous solutions. Manual intervention in the calibration allowed reliable parameters to be generated for all three missions.

  19. Ibis ground calibration

    International Nuclear Information System (INIS)

    Bird, A.J.; Barlow, E.J.; Tikkanen, T.; Bazzano, A.; Del Santo, M.; Ubertini, P.; Blondel, C.; Laurent, P.; Lebrun, F.; Di Cocco, G.; Malaguti, E.; Gabriele, M.; La Rosa, G.; Segreto, A.; Quadrini, E.; Volkmer, R.

    2003-01-01

    We present an overview of results obtained from IBIS ground calibrations. The spectral and spatial characteristics of the detector planes and surrounding passive materials have been determined through a series of calibration campaigns. Measurements of pixel gain, energy resolution, detection uniformity, efficiency and imaging capability are presented. The key results obtained from the ground calibration have been: - optimization of the instrument tunable parameters, - determination of energy linearity for all detection modes, - determination of energy resolution as a function of energy through the range 20 keV - 3 MeV, - demonstration of imaging capability in each mode, - measurement of intrinsic detector non-uniformity and understanding of the effects of passive materials surrounding the detector plane, and - discovery (and closure) of various leakage paths through the passive shielding system

  20. Multiple-Objective Stepwise Calibration Using Luca

    Science.gov (United States)

    Hay, Lauren E.; Umemoto, Makiko

    2007-01-01

    This report documents Luca (Let us calibrate), a multiple-objective, stepwise, automated procedure for hydrologic model calibration and the associated graphical user interface (GUI). Luca is a wizard-style user-friendly GUI that provides an easy systematic way of building and executing a calibration procedure. The calibration procedure uses the Shuffled Complex Evolution global search algorithm to calibrate any model compiled with the U.S. Geological Survey's Modular Modeling System. This process assures that intermediate and final states of the model are simulated consistently with measured values.

  1. Liquid Krypton Calorimeter Calibration Software

    CERN Document Server

    Hughes, Christina Lindsay

    2013-01-01

    Calibration of the liquid krypton calorimeter (LKr) of the NA62 experiment is managed by a set of standalone programs, or an online calibration driver. These programs are similar to those used by NA48, but have been updated to utilize classes and translated to C++ while maintaining a common functionality. A set of classes developed to handle communication with hardware was used to develop the three standalone programs as well as the main driver program for online calibration between bursts. The main calibration driver has been designed to respond to run control commands and receive burst data, both transmitted via DIM. In order to facilitate the process of reading in calibration parameters, a serializable class has been introduced, allowing the replacement of standard text files with XML configuration files.

  2. Compact radiometric microwave calibrator

    International Nuclear Information System (INIS)

    Fixsen, D. J.; Wollack, E. J.; Kogut, A.; Limon, M.; Mirel, P.; Singal, J.; Fixsen, S. M.

    2006-01-01

    The calibration methods for the ARCADE II instrument are described and the accuracy estimated. The Steelcast coated aluminum cones which comprise the calibrator have a low reflection while maintaining 94% of the absorber volume within 5 mK of the base temperature (modeled). The calibrator demonstrates an absorber with the active part less than one wavelength thick and only marginally larger than the mouth of the largest horn and yet black (less than -40 dB or 0.01% reflection) over five octaves in frequency

  3. Optimal, Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard

    2002-01-01

    Reliability based code calibration is considered in this paper. It is described how the results of FORM based reliability analysis may be related to the partial safety factors and characteristic values. The code calibration problem is presented in a decision theoretical form and it is discussed how...... of reliability based code calibration of LRFD based design codes....

  4. On the prospects of cross-calibrating the Cherenkov Telescope Array with an airborne calibration platform

    Science.gov (United States)

    Brown, Anthony M.

    2018-01-01

    Recent advances in unmanned aerial vehicle (UAV) technology have made UAVs an attractive possibility as an airborne calibration platform for astronomical facilities. This is especially true for arrays of telescopes spread over a large area such as the Cherenkov Telescope Array (CTA). In this paper, the feasibility of using UAVs to calibrate CTA is investigated. Assuming a UAV at 1km altitude above CTA, operating on astronomically clear nights with stratified, low atmospheric dust content, appropriate thermal protection for the calibration light source and an onboard photodiode to monitor its absolute light intensity, inter-calibration of CTA's telescopes of the same size class is found to be achievable with a 6 - 8 % uncertainty. For cross-calibration of different telescope size classes, a systematic uncertainty of 8 - 10 % is found to be achievable. Importantly, equipping the UAV with a multi-wavelength calibration light source affords us the ability to monitor the wavelength-dependent degradation of CTA telescopes' optical system, allowing us to not only maintain this 6 - 10 % uncertainty after the first few years of telescope deployment, but also to accurately account for the effect of multi-wavelength degradation on the cross-calibration of CTA by other techniques, namely with images of air showers and local muons. A UAV-based system thus provides CTA with several independent and complementary methods of cross-calibrating the optical throughput of individual telescopes. Furthermore, housing environmental sensors on the UAV system allows us to not only minimise the systematic uncertainty associated with the atmospheric transmission of the calibration signal, it also allows us to map the dust content above CTA as well as monitor the temperature, humidity and pressure profiles of the first kilometre of atmosphere above CTA with each UAV flight.

  5. Calibration method of liquid zone controller using the ex-core detector signal of CANDU 6 reactor

    International Nuclear Information System (INIS)

    Park, D.H.; Lee, E.K.; Shin, H.C.; Bae, S.M.; Hong, S.Y.

    2013-01-01

    Highlights: ► We developed a new LZC calibration method and measurement system. ► Photo-neutron effect, reactor core size, and detector position were evaluated and tested. ► We applied the new method and system to Wolsong NPP Unit 1. ► The LZC calibration test was well completed, and the requirement of the test was satisfied. - Abstract: The Phase-B test (low-power reactor physics test) is one of the commissioning tests for Canada Deuterium Uranium (CANDU) reactors that ensures the safe and reliable operation of the core during the design lifetime. The Phase-B test, which includes the approach to the first criticality at low reactor powers, is performed to verify the feasibility of the reactor’s physics design and to ensure the integrity of the control and protection facilities. The commissioning testing of pressurized heavy water moderated reactors (PHWRs) is usually performed only once (at the initial commissioning after construction). The large-scale facilities of the Wolsong nuclear power plant (NPP) Unit 1 have been gradually improved since May 2009 to extend its lifetime. The refurbishment was completed in April 2011 – then this NPP has been in operation again. We discusses the new methodology and measurement system that uses an ex-core detector signal for liquid zone controller (LZC) calibration of the Phase-B test instead of conventional methods. The inverse kinetic equation in the reactivity calculator is modified to treat the 17 delayed neutron groups including 11 photo-neutron fractions. The signal acquisition resolution of the reactivity calculator was enhanced and installed reactivity calculating module by each channel. The ex-core detector was confirmed to be applicable to a large reactor core, such as the CANDU 6 by comparison with the in-core flux detector signal. A preliminary test was performed in Wolsong NPP Unit 2 to verify the robustness of the reactivity calculator. This test convincingly demonstrated that the reactivity calculator

  6. A DXA Whole Body Composition Cross-Calibration Experience: Evaluation With Humans, Spine, and Whole Body Phantoms.

    Science.gov (United States)

    Krueger, Diane; Libber, Jessie; Sanfilippo, Jennifer; Yu, Hui Jing; Horvath, Blaine; Miller, Colin G; Binkley, Neil

    2016-01-01

    New densitometer installation requires cross-calibration for accurate longitudinal assessment. When replacing a unit with the same model, the International Society for Clinical Densitometry recommends cross-calibrating by scanning phantoms 10 times on each instrument and states that spine bone mineral density (BMD) should be within 1%, whereas total body lean, fat, and %fat mass should be within 2% of the prior instrument. However, there is limited validation that these recommendations provide adequate total body cross-calibration. Here, we report a total body cross-calibration experience with phantoms and humans. Cross-calibration between an existing and new Lunar iDXA was performed using 3 encapsulated spine phantoms (GE [GE Lunar, Madison, WI], BioClinica [BioClinica Inc, Princeton, NJ], and Hologic [Hologic Inc, Bedford, MA]), 1 total body composition phantom (BioClinica), and 30 human volunteers. Thirty scans of each phantom and a total body scan of human volunteers were obtained on each instrument. All spine phantom BMD means were similar (within 1%; g/cm2 bias) between the existing and new dual-energy X-ray absorptiometry unit. The BioClinica body composition phantom (BBCP) BMD and bone mineral content (BMC) values were within 2% with biases of 0.005 g/cm2 and -3.4 g. However, lean and fat mass and %fat differed by 4.6%-7.7% with biases of +463 g, -496 g, and -2.8%, respectively. In vivo comparison supported BBCP data; BMD and BMC were within ∼2%, but lean and fat mass and %fat differed from 1.6% to 4.9% with biases of +833 g, -860 g, and -1.1%. As all body composition comparisons exceeded the recommended 2%, the new densitometer was recalibrated. After recalibration, in vivo bias was lower (g, respectively. Similarly, BBCP lean and fat agreement improved. In conclusion, the BBCP behaves similarly, but not identical, to human in vivo measurements for densitometer cross-calibration. Spine phantoms, despite good BMD and BMC agreement, did not detect

  7. Calibration of higher eigenmodes of cantilevers

    International Nuclear Information System (INIS)

    Labuda, Aleksander; Kocun, Marta; Walsh, Tim; Meinhold, Jieh; Proksch, Tania; Meinhold, Waiman; Anderson, Caleb; Proksch, Roger; Lysy, Martin

    2016-01-01

    A method is presented for calibrating the higher eigenmodes (resonant modes) of atomic force microscopy cantilevers that can be performed prior to any tip-sample interaction. The method leverages recent efforts in accurately calibrating the first eigenmode by providing the higher-mode stiffness as a ratio to the first mode stiffness. A one-time calibration routine must be performed for every cantilever type to determine a power-law relationship between stiffness and frequency, which is then stored for future use on similar cantilevers. Then, future calibrations only require a measurement of the ratio of resonant frequencies and the stiffness of the first mode. This method is verified through stiffness measurements using three independent approaches: interferometric measurement, AC approach-curve calibration, and finite element analysis simulation. Power-law values for calibrating higher-mode stiffnesses are reported for several cantilever models. Once the higher-mode stiffnesses are known, the amplitude of each mode can also be calibrated from the thermal spectrum by application of the equipartition theorem.

  8. Calibration of higher eigenmodes of cantilevers

    Energy Technology Data Exchange (ETDEWEB)

    Labuda, Aleksander; Kocun, Marta; Walsh, Tim; Meinhold, Jieh; Proksch, Tania; Meinhold, Waiman; Anderson, Caleb; Proksch, Roger [Asylum Research, an Oxford Instruments Company, Santa Barbara, California 93117 (United States); Lysy, Martin [Department of Statistics and Actuarial Science, University of Waterloo, Waterloo, Ontario N2L 3G1 (Canada)

    2016-07-15

    A method is presented for calibrating the higher eigenmodes (resonant modes) of atomic force microscopy cantilevers that can be performed prior to any tip-sample interaction. The method leverages recent efforts in accurately calibrating the first eigenmode by providing the higher-mode stiffness as a ratio to the first mode stiffness. A one-time calibration routine must be performed for every cantilever type to determine a power-law relationship between stiffness and frequency, which is then stored for future use on similar cantilevers. Then, future calibrations only require a measurement of the ratio of resonant frequencies and the stiffness of the first mode. This method is verified through stiffness measurements using three independent approaches: interferometric measurement, AC approach-curve calibration, and finite element analysis simulation. Power-law values for calibrating higher-mode stiffnesses are reported for several cantilever models. Once the higher-mode stiffnesses are known, the amplitude of each mode can also be calibrated from the thermal spectrum by application of the equipartition theorem.

  9. SELECTION AND TREATMENT OF DATA FOR RADIOCARBON CALIBRATION : AN UPDATE TO THE INTERNATIONAL CALIBRATION (INTCAL) CRITERIA

    NARCIS (Netherlands)

    Reimer, Paula J.; Bard, Edouard; Bayliss, Alex; Beck, J. Warren; Blackwell, Paul G.; Ramsey, Christopher Bronk; Brown, David M.; Buck, Caitlin E.; Edwards, R. Lawrence; Friedrich, Michael; Grootes, Pieter M.; Guilderson, Thomas P.; Haflidason, Haflidi; Hajdas, Irka; Hatte, Christine; Heaton, Timothy J.; Hogg, Alan G.; Hughen, Konrad A.; Kaiser, K. Felix; Kromer, Bernd; Manning, Sturt W.; Reimer, Ron W.; Richards, David A.; Scott, E. Marian; Southon, John R.; Turney, Christian S. M.; van der Plicht, Johannes

    2013-01-01

    High-quality data from appropriate archives are needed for the continuing improvement of radiocarbon calibration curves. We discuss here the basic assumptions behind C-14 dating that necessitate calibration and the relative strengths and weaknesses of archives from which calibration data are

  10. A fundamental parameters approach to calibration of the Mars Exploration Rover Alpha Particle X-ray Spectrometer

    Science.gov (United States)

    Campbell, J. L.; Lee, M.; Jones, B. N.; Andrushenko, S. M.; Holmes, N. G.; Maxwell, J. A.; Taylor, S. M.

    2009-04-01

    The detection sensitivities of the Alpha Particle X-ray Spectrometer (APXS) instruments on the Mars Exploration Rovers for a wide range of elements were experimentally determined in 2002 using spectra of geochemical reference materials. A flight spare instrument was similarly calibrated, and the calibration exercise was then continued for this unit with an extended set of geochemical reference materials together with pure elements and simple chemical compounds. The flight spare instrument data are examined in detail here using a newly developed fundamental parameters approach which takes precise account of all the physics inherent in the two X-ray generation techniques involved, namely, X-ray fluorescence and particle-induced X-ray emission. The objectives are to characterize the instrument as fully as possible, to test this new approach, and to determine the accuracy of calibration for major, minor, and trace elements. For some of the lightest elements the resulting calibration exhibits a dependence upon the mineral assemblage of the geological reference material; explanations are suggested for these observations. The results will assist in designing the overall calibration approach for the APXS on the Mars Science Laboratory mission.

  11. Calibration technique and study on metrological characteristics of a high-voltage inverse square-law function generator

    International Nuclear Information System (INIS)

    Popov, V.P.; Semenov, A.L.

    1987-01-01

    The calibration technique is described, and the metrological characteristics of a high-voltage generator of the inverse-quadratic function (HGF), being a functional unit of the diagnostic system of an electrodynamic analyser of a ionic component of a laser plasma, is analysed. The results of HGF testing in the range of time constants of the τ=(5-25)μs function are given. Analysis of metrologic and experimental characteristics shows, that HGF with automatic calibration has quite high accurate parameters. The high accuracy of function generation is provided with the possibility of calibration and adjustment conduction under experimental working conditions. Increase of the generated pulse amplitude to several tens of kilovelts is possible. Besides, the possibility of timely function adjustment to the necessary parameter (τ) increases essentially the HGF functional possibilities

  12. Calibration artefacts in radio interferometry - III. Phase-only calibration and primary beam correction

    Science.gov (United States)

    Grobler, T. L.; Stewart, A. J.; Wijnholds, S. J.; Kenyon, J. S.; Smirnov, O. M.

    2016-09-01

    This is the third installment in a series of papers in which we investigate calibration artefacts. Calibration artefacts (also known as ghosts or spurious sources) are created when we calibrate with an incomplete model. In the first two papers of this series, we developed a mathematical framework which enabled us to study the ghosting mechanism itself. An interesting concomitant of the second paper was that ghosts appear in symmetrical pairs. This could possibly account for spurious symmetrization. Spurious symmetrization refers to the appearance of a spurious source (the antighost) symmetrically opposite an unmodelled source around a modelled source. The analysis in the first two papers indicates that the antighost is usually very faint, in particular, when a large number of antennas are used. This suggests that spurious symmetrization will mainly occur at an almost undetectable flux level. In this paper, we show that phase-only calibration produces an antighost that is N-times (where N denotes the number of antennas in the array) as bright as the one produced by phase and amplitude calibration and that this already bright ghost can be further amplified by the primary beam correction.

  13. In-situ calibration: migrating control system IP module calibration from the bench to the storage ring

    International Nuclear Information System (INIS)

    Weber, Jonah M.; Chin, Michael

    2002-01-01

    The Control System for the Advanced Light Source (ALS) at Lawrence Berkeley National Lab (LBNL) uses in-house designed IndustryPack(registered trademark) (IP) modules contained in compact PCI (cPCI) crates with 16-bit analog I/O to control instrumentation. To make the IP modules interchangeable, each module is calibrated for gain and offset compensation. We initially developed a method of verifying and calibrating the IP modules in a lab bench test environment using a PC with LabVIEW. The subsequent discovery that the ADCs have significant drift characteristics over periods of days of installed operation prompted development of an ''in-situ'' calibration process--one in which the IP modules can be calibrated without removing them from the cPCI crates in the storage ring. This paper discusses the original LabVIEW PC calibration and the migration to the proposed in-situ EPICS control system calibration

  14. New thermal neutron calibration channel at LNMRI/IRD

    International Nuclear Information System (INIS)

    Astuto, A.; Lopes, R.T.; Patrao, K.C.S.; Fonseca, E.S.; Pereira, W.W.

    2015-01-01

    A new standard thermal neutron flux unit was designed in the National Ionizing Radiation Metrology Laboratory (LNMRI) for calibration of neutron detectors. Fluence is achieved by moderation of four 241 Am-Be sources with 596 GBq each, in a facility built with graphite and paraffin blocks. The study was divided into two stages. First, simulations were performed using MCNPX code in different geometric arrangements, seeking the best performance in terms of fluence and their uncertainties. Last, the system was assembled based on the results obtained on the simulations. The simulation results indicate quasi-homogeneous fluence (less than 1%) in the central chamber. (author)

  15. Technical guidelines for personnel dosimetry calibrations

    International Nuclear Information System (INIS)

    Roberson, P.L.; Fox, R.A.; Hadley, R.T.; Holbrook, K.L.; Hooker, C.D.; McDonald, J.C.

    1983-01-01

    A base of technical information has been acquire and used to evaluate the calibration, design, and performance of selected personnel systems in use at Department of Energy (DOE) facilites. A technical document was prepared to guide DOE and DOE contractors in selecting and evaluating personnel dosimetry systems and calibration. A parallel effort was initiated to intercompare the adiological calibrations standards used to calibrate DOE personnel dosimeters

  16. Experimental comparison between total calibration factors and components calibration factors of reference dosemeters used in secondary standard laboratory dosemeters

    International Nuclear Information System (INIS)

    Silva, T.A. da.

    1981-06-01

    A quantitative comparison of component calibration factors with the corresponding overall calibration factor was used to evaluate the adopted component calibration procedure in regard to parasitic elements. Judgement of significance is based upon the experimental uncertainty of a well established procedure for determination of the overall calibration factor. The experimental results obtained for different ionization chambers and different electrometers demonstrate that for one type of electrometer the parasitic elements have no influence on its sensitivity considering the experimental uncertainty of the calibration procedures. In this case the adopted procedure for determination of component calibration factors is considered to be equivalent to the procedure of determination of the overall calibration factor and thus might be used as a strong quality control measure in routine calibration. (Author) [pt

  17. Calibration of personal dosemeters in terms of the ICRU operational quantities

    International Nuclear Information System (INIS)

    McDonald, J.C.; Hertel, N.E.

    1992-01-01

    The International Commission on Radiological Units and Measurements (ICRU) has defined several new operational quantities for radiation protection purposes. The quantities to be used for personal monitoring are defined at depths in the human body. Because these quantities are impossible to measure directly, the ICRU has recommended that personal dosimeters should be calibrated under simplified conditions on an appropriate phantom, such as the ICRU sphere. The U.S. personal dosimetry accreditation programs make use of a 30 x 30 x 15 cm polymethylmethacrylate (PMMA) phantom; therefore it is necessary to relate the response of dosimeters calibrated on this phantom to the ICRU operational quantities. Calculations of the conversion factors to compute dosimeter response in terms of the operational quantities have been performed using the code MCNP. These calculations have also been compared to experimental measurements using thermoluminescent (TLD) detectors. (author)

  18. Calibration of personal dosemeters in terms of the ICRU operational quantities

    International Nuclear Information System (INIS)

    McDonald, J.C.; Hertel, N.E.

    1992-05-01

    The International Commission on Radiological Units and Measurements (ICRU) has defined several new operational quantities for radiation protection purposes. The quantities to be used for personal monitoring are defined at depths in the human body. Because these quantities are impossible to measure directly, the ICRU has recommended that personal dosemeters should be calibrated under simplified conditions on an appropriate phantom, such as the ICRU sphere. The US personal dosimetry accreditation programs make use of a 30 x 30 x 15 cm polymethymethacrylate (PMMA) phantom, therefore it is necessary to relate the response of dosemeters calibrated on this phantom to the ICRU operational quantities. Calculations of the conversion factors to compute dosemeter response in terms of the operational quantities have been performed using the code MCNP. These calculations have also been compared to experimental measurements using thermoluminescent (TLD) detectors

  19. Iterative Calibration: A Novel Approach for Calibrating the Molecular Clock Using Complex Geological Events.

    Science.gov (United States)

    Loeza-Quintana, Tzitziki; Adamowicz, Sarah J

    2018-02-01

    During the past 50 years, the molecular clock has become one of the main tools for providing a time scale for the history of life. In the era of robust molecular evolutionary analysis, clock calibration is still one of the most basic steps needing attention. When fossil records are limited, well-dated geological events are the main resource for calibration. However, biogeographic calibrations have often been used in a simplistic manner, for example assuming simultaneous vicariant divergence of multiple sister lineages. Here, we propose a novel iterative calibration approach to define the most appropriate calibration date by seeking congruence between the dates assigned to multiple allopatric divergences and the geological history. Exploring patterns of molecular divergence in 16 trans-Bering sister clades of echinoderms, we demonstrate that the iterative calibration is predominantly advantageous when using complex geological or climatological events-such as the opening/reclosure of the Bering Strait-providing a powerful tool for clock dating that can be applied to other biogeographic calibration systems and further taxa. Using Bayesian analysis, we observed that evolutionary rate variability in the COI-5P gene is generally distributed in a clock-like fashion for Northern echinoderms. The results reveal a large range of genetic divergences, consistent with multiple pulses of trans-Bering migrations. A resulting rate of 2.8% pairwise Kimura-2-parameter sequence divergence per million years is suggested for the COI-5P gene in Northern echinoderms. Given that molecular rates may vary across latitudes and taxa, this study provides a new context for dating the evolutionary history of Arctic marine life.

  20. Solid laboratory calibration of a nonimaging spectroradiometer.

    Science.gov (United States)

    Schaepman, M E; Dangel, S

    2000-07-20

    Field-based nonimaging spectroradiometers are often used in vicarious calibration experiments for airborne or spaceborne imaging spectrometers. The calibration uncertainties associated with these ground measurements contribute substantially to the overall modeling error in radiance- or reflectance-based vicarious calibration experiments. Because of limitations in the radiometric stability of compact field spectroradiometers, vicarious calibration experiments are based primarily on reflectance measurements rather than on radiance measurements. To characterize the overall uncertainty of radiance-based approaches and assess the sources of uncertainty, we carried out a full laboratory calibration. This laboratory calibration of a nonimaging spectroradiometer is based on a measurement plan targeted at achieving a calibration. The individual calibration steps include characterization of the signal-to-noise ratio, the noise equivalent signal, the dark current, the wavelength calibration, the spectral sampling interval, the nonlinearity, directional and positional effects, the spectral scattering, the field of view, the polarization, the size-of-source effects, and the temperature dependence of a particular instrument. The traceability of the radiance calibration is established to a secondary National Institute of Standards and Technology calibration standard by use of a 95% confidence interval and results in an uncertainty of less than ?7.1% for all spectroradiometer bands.

  1. Muon Energy Calibration of the MINOS Detectors

    Energy Technology Data Exchange (ETDEWEB)

    Miyagawa, Paul S. [Somerville College, Oxford (United Kingdom)

    2004-01-01

    MINOS is a long-baseline neutrino oscillation experiment designed to search for conclusive evidence of neutrino oscillations and to measure the oscillation parameters precisely. MINOS comprises two iron tracking calorimeters located at Fermilab and Soudan. The Calibration Detector at CERN is a third MINOS detector used as part of the detector response calibration programme. A correct energy calibration between these detectors is crucial for the accurate measurement of oscillation parameters. This thesis presents a calibration developed to produce a uniform response within a detector using cosmic muons. Reconstruction of tracks in cosmic ray data is discussed. This data is utilized to calculate calibration constants for each readout channel of the Calibration Detector. These constants have an average statistical error of 1.8%. The consistency of the constants is demonstrated both within a single run and between runs separated by a few days. Results are presented from applying the calibration to test beam particles measured by the Calibration Detector. The responses are calibrated to within 1.8% systematic error. The potential impact of the calibration on the measurement of oscillation parameters by MINOS is also investigated. Applying the calibration reduces the errors in the measured parameters by ~ 10%, which is equivalent to increasing the amount of data by 20%.

  2. The KLOE online calibration system

    International Nuclear Information System (INIS)

    Pasqualucci, E.; Alexander, G.; Aloisio, A.

    2001-01-01

    Based on all the features of the KLOE online software, the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed. A calibration manager process controls the system, implementing the interface to the online system, receiving information from the run control and translating its state transitions to a separate state machine. It acts as a 'calibration run controller' and performs failure recovery when requested by a set of process checkers. The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms. A client library and C, fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an book-like interface. Several calibration processes running in parallel in a distributed, multiplatform environment can fill the same histograms, allowing fast external information check. A monitor thread allow remote browsing for visual inspection. Pre-filtered data are read in non-privileged spy mode from the data acquisition system via the Kloe Integrated Dataflow. The main characteristics of the system are presented

  3. Calibration-free optical chemical sensors

    Science.gov (United States)

    DeGrandpre, Michael D.

    2006-04-11

    An apparatus and method for taking absorbance-based chemical measurements are described. In a specific embodiment, an indicator-based pCO2 (partial pressure of CO2) sensor displays sensor-to-sensor reproducibility and measurement stability. These qualities are achieved by: 1) renewing the sensing solution, 2) allowing the sensing solution to reach equilibrium with the analyte, and 3) calculating the response from a ratio of the indicator solution absorbances which are determined relative to a blank solution. Careful solution preparation, wavelength calibration, and stray light rejection also contribute to this calibration-free system. Three pCO2 sensors were calibrated and each had response curves which were essentially identical within the uncertainty of the calibration. Long-term laboratory and field studies showed the response had no drift over extended periods (months). The theoretical response, determined from thermodynamic characterization of the indicator solution, also predicted the observed calibration-free performance.

  4. Evaluation of plasmid and genomic DNA calibrants used for the quantification of genetically modified organisms.

    Science.gov (United States)

    Caprioara-Buda, M; Meyer, W; Jeynov, B; Corbisier, P; Trapmann, S; Emons, H

    2012-07-01

    The reliable quantification of genetically modified organisms (GMOs) by real-time PCR requires, besides thoroughly validated quantitative detection methods, sustainable calibration systems. The latter establishes the anchor points for the measured value and the measurement unit, respectively. In this paper, the suitability of two types of DNA calibrants, i.e. plasmid DNA and genomic DNA extracted from plant leaves, for the certification of the GMO content in reference materials as copy number ratio between two targeted DNA sequences was investigated. The PCR efficiencies and coefficients of determination of the calibration curves as well as the measured copy number ratios for three powder certified reference materials (CRMs), namely ERM-BF415e (NK603 maize), ERM-BF425c (356043 soya), and ERM-BF427c (98140 maize), originally certified for their mass fraction of GMO, were compared for both types of calibrants. In all three systems investigated, the PCR efficiencies of plasmid DNA were slightly closer to the PCR efficiencies observed for the genomic DNA extracted from seed powders rather than those of the genomic DNA extracted from leaves. Although the mean DNA copy number ratios for each CRM overlapped within their uncertainties, the DNA copy number ratios were significantly different using the two types of calibrants. Based on these observations, both plasmid and leaf genomic DNA calibrants would be technically suitable as anchor points for the calibration of the real-time PCR methods applied in this study. However, the most suitable approach to establish a sustainable traceability chain is to fix a reference system based on plasmid DNA.

  5. NEW METHOD FOR THE CALIBRATION OF MULTI-CAMERA MOBILE MAPPING SYSTEMS

    Directory of Open Access Journals (Sweden)

    A. P. Kersting

    2012-07-01

    Full Text Available Mobile Mapping Systems (MMS allow for fast and cost-effective collection of geo-spatial information. Such systems integrate a set of imaging sensors and a position and orientation system (POS, which entails GPS and INS units. System calibration is a crucial process to ensure the attainment of the expected accuracy of such systems. It involves the calibration of the individual sensors as well as the calibration of the mounting parameters relating the system components. The mounting parameters of multi-camera MMS include two sets of relative orientation parameters (ROP: the lever arm offsets and the boresight angles relating the cameras and the IMU body frame and the ROP among the cameras (in the absence of GPS/INS data. In this paper, a novel single-step calibration method, which has the ability of estimating these two sets of ROP, is devised. Besides the ability to estimate the ROP among the cameras, the proposed method can use such parameters as prior information in the ISO procedure. The implemented procedure consists of an integrated sensor orientation (ISO where the GPS/INS-derived position and orientation and the system mounting parameters are directly incorporated in the collinearity equations. The concept of modified collinearity equations has been used by few authors for single-camera systems. In this paper, a new modification to the collinearity equations for GPS/INS-assisted multicamera systems is introduced. Experimental results using a real dataset demonstrate the feasibility of the proposed method.

  6. New Method for the Calibration of Multi-Camera Mobile Mapping Systems

    Science.gov (United States)

    Kersting, A. P.; Habib, A.; Rau, J.

    2012-07-01

    Mobile Mapping Systems (MMS) allow for fast and cost-effective collection of geo-spatial information. Such systems integrate a set of imaging sensors and a position and orientation system (POS), which entails GPS and INS units. System calibration is a crucial process to ensure the attainment of the expected accuracy of such systems. It involves the calibration of the individual sensors as well as the calibration of the mounting parameters relating the system components. The mounting parameters of multi-camera MMS include two sets of relative orientation parameters (ROP): the lever arm offsets and the boresight angles relating the cameras and the IMU body frame and the ROP among the cameras (in the absence of GPS/INS data). In this paper, a novel single-step calibration method, which has the ability of estimating these two sets of ROP, is devised. Besides the ability to estimate the ROP among the cameras, the proposed method can use such parameters as prior information in the ISO procedure. The implemented procedure consists of an integrated sensor orientation (ISO) where the GPS/INS-derived position and orientation and the system mounting parameters are directly incorporated in the collinearity equations. The concept of modified collinearity equations has been used by few authors for single-camera systems. In this paper, a new modification to the collinearity equations for GPS/INS-assisted multicamera systems is introduced. Experimental results using a real dataset demonstrate the feasibility of the proposed method.

  7. Calibration and energy response of the Bitt RM10/RS02 gamma radiation detectors

    International Nuclear Information System (INIS)

    Dijk, E. van; Aalbers, A.H.L.

    1990-03-01

    A radiation monitoring network with automatic warning capabilities (LMR) has been established in the Netherlands. For the detection of gamma radiation exposure-rate-meters manufactured by Bitt Technologies are used. These meters consist of a proportional counter tube (type RS 02) and a read-out unit (type RM 10E). The photon energy response of 6 counter tubes was tested at the National Institute of Public Health and Environmental Protection. The measurements were performed with heavy filtered X-rays in the range of 50-250 keV (ISO narrow spectrum series) and with gamma ray beams from cesium-137 (662 keV) and cobalt-60 (1,25 MeV). To determine the energy response, the detector reading was referred to air kerma by means of a transfer ionization chamber. This transfer chamber was directly calibrated against the standard for X-rays. By applying these measurement procedures of a set of calibration factors (N k ) as a function of photon energy was determined. These calibration factors, expressed as the ratio air kerma to reading were converted to ambient dose equivalent calibration factors using appropriate conversion factors taken from Grosswend et al., 1988. From the measurement data an average ambient dose equivalent calibration factor of 10.8 mSv.roentgen -1 was calculated. (author). 5 refs.; 6 figs.; 5 tabs

  8. Calorimetry of energy-efficient glow discharge apparatus design and calibration

    International Nuclear Information System (INIS)

    Benson, Thomas B.; Passell, Thomas O.

    2006-01-01

    This work aims to develop a 'family' or low-powered calorimetrically accurate glow discharge units, similar to that reported by Dardik et al. at lCCF-10, and to use these to test a wide range or cathode materials, electrode coatings, gas types, gas pressures, and power input levels. We will describe the design and calibration of these units. The strategy is to use a large number of very similar units so that the calorimetric response does not vary significantly for a given power level. The design is metal or sealed glass cylindrical tubes, charged with 0.4 - 50 Torr mixtures of deuterium, hydrogen, argon, or helium gases. Units operate from 0.2 to >2 W power input. The units have low mass ( 1.2 with more than 95% certainty. It provides a valuable new platform for large-scale exploration of excess heat effects in the gas phase, using low-power inputs in the 0-3 W range, This method proves to be inexpensive, quick, accurate, and easy to perform once the basics are mastered. The authors are interested in testing electrode materials from other sources, especially those that have already been successful in a liquid (electrolytic) environment

  9. The Chandra X-ray Observatory PSF Library

    Science.gov (United States)

    Karovska, M.; Beikman, S. J.; Elvis, M. S.; Flanagan, J. M.; Gaetz, T.; Glotfelty, K. J.; Jerius, D.; McDowell, J. C.; Rots, A. H.

    Pre-flight and on-orbit calibration of the Chandra X-Ray Observatory provided a unique base for developing detailed models of the optics and detectors. Using these models we have produced a set of simulations of the Chandra point spread function (PSF) which is available to the users via PSF library files. We describe here how the PSF models are generated and the design and content of the Chandra PSF library files.

  10. Error-in-variables models in calibration

    Science.gov (United States)

    Lira, I.; Grientschnig, D.

    2017-12-01

    In many calibration operations, the stimuli applied to the measuring system or instrument under test are derived from measurement standards whose values may be considered to be perfectly known. In that case, it is assumed that calibration uncertainty arises solely from inexact measurement of the responses, from imperfect control of the calibration process and from the possible inaccuracy of the calibration model. However, the premise that the stimuli are completely known is never strictly fulfilled and in some instances it may be grossly inadequate. Then, error-in-variables (EIV) regression models have to be employed. In metrology, these models have been approached mostly from the frequentist perspective. In contrast, not much guidance is available on their Bayesian analysis. In this paper, we first present a brief summary of the conventional statistical techniques that have been developed to deal with EIV models in calibration. We then proceed to discuss the alternative Bayesian framework under some simplifying assumptions. Through a detailed example about the calibration of an instrument for measuring flow rates, we provide advice on how the user of the calibration function should employ the latter framework for inferring the stimulus acting on the calibrated device when, in use, a certain response is measured.

  11. Calibration Errors in Interferometric Radio Polarimetry

    Science.gov (United States)

    Hales, Christopher A.

    2017-08-01

    Residual calibration errors are difficult to predict in interferometric radio polarimetry because they depend on the observational calibration strategy employed, encompassing the Stokes vector of the calibrator and parallactic angle coverage. This work presents analytic derivations and simulations that enable examination of residual on-axis instrumental leakage and position-angle errors for a suite of calibration strategies. The focus is on arrays comprising alt-azimuth antennas with common feeds over which parallactic angle is approximately uniform. The results indicate that calibration schemes requiring parallactic angle coverage in the linear feed basis (e.g., the Atacama Large Millimeter/submillimeter Array) need only observe over 30°, beyond which no significant improvements in calibration accuracy are obtained. In the circular feed basis (e.g., the Very Large Array above 1 GHz), 30° is also appropriate when the Stokes vector of the leakage calibrator is known a priori, but this rises to 90° when the Stokes vector is unknown. These findings illustrate and quantify concepts that were previously obscure rules of thumb.

  12. Calibration strategies for the Cherenkov Telescope Array

    NARCIS (Netherlands)

    Gaug, M.; Berge, D.; Daniel, M.; Doro, M.; Förster, A.; Hofmann, W.; Maccarone, M.C.; Parsons, D.; de los Reyes Lopez, R.; van Eldik, C.

    2014-01-01

    The Central Calibration Facilities workpackage of the Cherenkov Telescope Array (CTA) observatory for very high energy gamma ray astronomy defines the overall calibration strategy of the array, develops dedicated hardware and software for the overall array calibration and coordinates the calibration

  13. Cross-calibration of interferometric SAR data

    DEFF Research Database (Denmark)

    Dall, Jørgen

    2003-01-01

    Generation of digital elevation models from interferometric synthetic aperture radar (SAR) data is a well established technique. Achieving a high geometric fidelity calls for a calibration accounting for inaccurate navigation data and system parameters as well as system imperfections. Fully...... automated calibration techniques are preferable, especially for operational mapping. The author presents one such technique, called cross-calibration. Though developed for single-pass interferometry, it may be applicable to multi-pass interferometry, too. Cross-calibration requires stability during mapping...... ground control point is often needed. The paper presents the principles and mathematics of the cross-calibration technique and illustrates its successful application to EMISAR data....

  14. Calibration of nuclear medicine gamma counters

    International Nuclear Information System (INIS)

    Orlic, M.; Spasic-Jokic, V.; Jovanovic, M.; Vranjes, S. . E-mail address of corresponding author: morlic@vin.bg.ac.yu; Orlic, M.)

    2005-01-01

    In this paper the practical problem of nuclear medicine gamma counters calibration has been solved by using dose calibrators CRC-15R with standard error ±5%. The samples from technetium generators have been measured both by dose calibrators CRC-15R and gamma counter ICN Gamma 3.33 taking into account decay correction. Only the linear part of the curve has practical meaning. The advantage of this procedure satisfies the requirements from international standards: the calibration of sources used for medical exposure be traceable to a standard dosimetry laboratory and radiopharmaceuticals for nuclear medicine procedures be calibrated in terms of activity of the radiopharmaceutical to be administered. (author)

  15. Automatic calibration of gamma spectrometers

    International Nuclear Information System (INIS)

    Tluchor, D.; Jiranek, V.

    1989-01-01

    The principle is described of energy calibration of the spectrometric path based on the measurement of the standard of one radionuclide or a set of them. The entire computer-aided process is divided into three main steps, viz.: the insertion of the calibration standard by the operator; the start of the calibration program; energy calibration by the computer. The program was selected such that the spectrum identification should not depend on adjustment of the digital or analog elements of the gamma spectrometric measuring path. The ECL program is described for automatic energy calibration as is its control, the organization of data file ECL.DAT and the necessary hardware support. The computer-multichannel analyzer communication was provided using an interface pair of Canberra 8673V and Canberra 8573 operating in the RS-422 standard. All subroutines for communication with the multichannel analyzer were written in MACRO 11 while the main program and the other subroutines were written in FORTRAN-77. (E.J.). 1 tab., 4 refs

  16. Permanently calibrated interpolating time counter

    International Nuclear Information System (INIS)

    Jachna, Z; Szplet, R; Kwiatkowski, P; Różyc, K

    2015-01-01

    We propose a new architecture of an integrated time interval counter that provides its permanent calibration in the background. Time interval measurement and the calibration procedure are based on the use of a two-stage interpolation method and parallel processing of measurement and calibration data. The parallel processing is achieved by a doubling of two-stage interpolators in measurement channels of the counter, and by an appropriate extension of control logic. Such modification allows the updating of transfer characteristics of interpolators without the need to break a theoretically infinite measurement session. We describe the principle of permanent calibration, its implementation and influence on the quality of the counter. The precision of the presented counter is kept at a constant level (below 20 ps) despite significant changes in the ambient temperature (from −10 to 60 °C), which can cause a sevenfold decrease in the precision of the counter with a traditional calibration procedure. (paper)

  17. Comparison of the radiochromic EBT2 responses for 4MV LINACs in calibration processes

    Energy Technology Data Exchange (ETDEWEB)

    Silva, Sabrina D.; Castro, André L.S.; Mendes, Thais M.; Campos, Tarcisio P.R. [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Dept. de Energia Nuclear; Nogueira, Luciana B., E-mail: sadonatosilva@hotmail.com, E-mail: radioterapia.andre@gmail.com, E-mail: thaismelomendes@yahoo.com.br, E-mail: tprcampos@yahoo.com.br, E-mail: Lucibn19@yahoo.com.br [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Anatomia e Imagem

    2017-07-01

    Background: Actually, cancer has gained a larger dimension and become a global public health problem. Radiotherapy (RT) is a neoplasia treatment and RT linear accelerators must undergo a strict dose quality control. Pure liquid water or solid water phantoms can be used with this intuit. In recent years, radiochromic films with equivalent tissue composition have been widely used as dosimeters in the medical field. Objective: the proposal was to analyze two distinct radiochromic film responses in water and solid water phantoms, in a distinct LINAC spectra of 4MV beam. Methods: Solid water phantom and EBT2 Radiochromic films were set. Films were exposed to a set of absorbed doses established by distinct monitor units (MU) specified in each RT-center. Mathematical relations between the degree of red-intensity from digitized films and the absorbed dose for both spectra were established. The coefficients of the polynomial function of the calibration curve were determined from the Origin software. The uncertainty of both processes was analyzed. The efficiency of the two calibration processes was set up. Results: The adjustment of the calibration curve provided the coefficients of the second-order equation that relates the dose absorbed with the optical density of the film. The uncertainty regarding the calibration performed in water and solid water and the dose-error accuracy are in agreement with the literature. Conclusion: Both water or solid water were effective in calibration and can be used in routines of quality-control measurements. The results show that EBT2-radiochromic films is suitable to for dose-calibration in RT. (author)

  18. Pre-flight calibration and initial data processing for the ChemCam laser-induced breakdown spectroscopy instrument on the Mars Science Laboratory rover

    Science.gov (United States)

    Wiens, R.C.; Maurice, S.; Lasue, J.; Forni, O.; Anderson, R.B.; Clegg, S.; Bender, S.; Blaney, D.; Barraclough, B.L.; Cousin, A.; DeFlores, L.; Delapp, D.; Dyar, M.D.; Fabre, C.; Gasnault, O.; Lanza, N.; Mazoyer, J.; Melikechi, N.; Meslin, P.-Y.; Newsom, H.; Ollila, A.; Perez, R.; Tokar, R.; Vaniman, D.

    2013-01-01

    The ChemCam instrument package on the Mars Science Laboratory rover, Curiosity, is the first planetary science instrument to employ laser-induced breakdown spectroscopy (LIBS) to determine the compositions of geological samples on another planet. Pre-processing of the spectra involves subtracting the ambient light background, removing noise, removing the electron continuum, calibrating for the wavelength, correcting for the variable distance to the target, and applying a wavelength-dependent correction for the instrument response. Further processing of the data uses multivariate and univariate comparisons with a LIBS spectral library developed prior to launch as well as comparisons with several on-board standards post-landing. The level-2 data products include semi-quantitative abundances derived from partial least squares regression. A LIBS spectral library was developed using 69 rock standards in the form of pressed powder disks, glasses, and ceramics to minimize heterogeneity on the scale of the observation (350–550 μm dia.). The standards covered typical compositional ranges of igneous materials and also included sulfates, carbonates, and phyllosilicates. The provenance and elemental and mineralogical compositions of these standards are described. Spectral characteristics of this data set are presented, including the size distribution and integrated irradiances of the plasmas, and a proxy for plasma temperature as a function of distance from the instrument. Two laboratory-based clones of ChemCam reside in Los Alamos and Toulouse for the purpose of adding new spectra to the database as the need arises. Sensitivity to differences in wavelength correlation to spectral channels and spectral resolution has been investigated, indicating that spectral registration needs to be within half a pixel and resolution needs to match within 1.5 to 2.6 pixels. Absolute errors are tabulated for derived compositions of each major element in each standard using PLS regression

  19. Radiometric calibration of the in-flight blackbody calibration system of the GLORIA interferometer

    Directory of Open Access Journals (Sweden)

    C. Monte

    2014-01-01

    Atmosphere is an airborne, imaging, infrared Fourier transform spectrometer that applies the limb-imaging technique to perform trace gas and temperature measurements in the Earth's atmosphere with three-dimensional resolution. To ensure the traceability of these measurements to the International Temperature Scale and thereby to an absolute radiance scale, GLORIA carries an on-board calibration system. Basically, it consists of two identical large-area and high-emissivity infrared radiators, which can be continuously and independently operated at two adjustable temperatures in a range from −50 °C to 0 °C during flight. Here we describe the radiometric and thermometric characterization and calibration of the in-flight calibration system at the Reduced Background Calibration Facility of the Physikalisch-Technische Bundesanstalt. This was performed with a standard uncertainty of less than 110 mK. Extensive investigations of the system concerning its absolute radiation temperature and spectral radiance, its temperature homogeneity and its short- and long-term stability are discussed. The traceability chain of these measurements is presented.

  20. e-Calibrations: using the Internet to deliver calibration services in real time at lower cost

    International Nuclear Information System (INIS)

    Desrosiers, Marc; Nagy, Vitaly; Puhl, James; Glenn, Robert; Densock, Robert; Stieren, David; Lang, Brian; Kamlowski, Andreas; Maier, Diether; Heiss, Arthur

    2002-01-01

    The National Institute of Standards and Technology (NIST) is expanding into a new frontier in the delivery of measurement services. The Internet will be employed to provide industry with electronic traceability to national standards. This is a radical departure from the traditional modes of traceability and presents many new challenges. The traditional mail-based calibration service relies on sending artifacts to the user, who then mails them back to NIST for evaluation. The new service will deliver calibration results to the industry customer on-demand, in real-time, at a lower cost. The calibration results can be incorporated rapidly into the production process to ensure the highest quality manufacturing. The service would provide the US radiation processing industry with a direct link to the NIST calibration facilities and its expertise, and provide an interactive feedback process between industrial processing and the national measurement standard. Moreover, an Internet calibration system should contribute to the removal of measurement-related trade barriers

  1. Automated Calibration of Dosimeters for Diagnostic Radiology

    International Nuclear Information System (INIS)

    Romero Acosta, A.; Gutierrez Lores, S.

    2015-01-01

    Calibration of dosimeters for diagnostic radiology includes current and charge measurements, which are often repetitive. However, these measurements are usually done using modern electrometers, which are equipped with an RS-232 interface that enables instrument control from a computer. This paper presents an automated system aimed to the measurements for the calibration of dosimeters used in diagnostic radiology. A software application was developed, in order to achieve the acquisition of the electric charge readings, measured values of the monitor chamber, calculation of the calibration coefficient and issue of a calibration certificate. A primary data record file is filled and stored in the computer hard disk. The calibration method used was calibration by substitution. With this system, a better control over the calibration process is achieved and the need for human intervention is reduced. the automated system will be used in the calibration of dosimeters for diagnostic radiology at the Cuban Secondary Standard Dosimetry Laboratory of the Center for Radiation Protection and Hygiene. (Author)

  2. X-ray calibration qualities

    International Nuclear Information System (INIS)

    Burns, J.E.

    1998-01-01

    Since the recent publication of IAEA Technical Reports Series No. 374 ''Calibration of Dosimeters Used in Radiotheraphy'', there have been a number of queries about the origin of, and the rationale behind, the X-ray qualities recommended for calibration purposes. The simple answer is that these are the qualities derived at the UK National Physical Laboratory (NPL) in 1971 for calibration of therapy-level dosimeters and which are still in use for that purpose. As some SSDLs may have difficulties in adopting these exact combinations of kV and filtration. This paper discusses the basic ideas involved, and how to go about deriving a different series of qualities

  3. 40 CFR 1065.310 - Torque calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 32 2010-07-01 2010-07-01 false Torque calibration. 1065.310 Section... Conditions § 1065.310 Torque calibration. (a) Scope and frequency. Calibrate all torque-measurement systems including dynamometer torque measurement transducers and systems upon initial installation and after major...

  4. Observatory Magnetometer In-Situ Calibration

    Directory of Open Access Journals (Sweden)

    A Marusenkov

    2011-07-01

    Full Text Available An experimental validation of the in-situ calibration procedure, which allows estimating parameters of observatory magnetometers (scale factors, sensor misalignment without its operation interruption, is presented. In order to control the validity of the procedure, the records provided by two magnetometers calibrated independently in a coil system have been processed. The in-situ estimations of the parameters are in very good agreement with the values provided by the coil system calibration.

  5. Calibration of high-resolution electronic autocollimators with demanded low uncertainties using single reading head angle encoders

    International Nuclear Information System (INIS)

    Yandayan, Tanfer; Akgoz, S Asli; Asar, Muharrem

    2014-01-01

    Calibration of high-resolution electronic autocollimators is carried out in TUBITAK UME using an angle comparator to ensure direct traceability to the SI unit of plane angle, radian (rad). The device is a specially designed air-bearing rotary table fitted with a commercially available angular encoder utilizing a single reading head. It is shown that high-resolution electronic autocollimators in the large measurement range (e.g. ±1000 arcsec) can be calibrated with an expanded uncertainty of 0.035 arcsec (k = 2) in conventional dimensional laboratory conditions, applying good measurement strategy for single reading head angle encoders and taking simple but smart precautions. Description of the angle comparator is presented with various test results derived using different high-precision autocollimators, and a detailed uncertainty budget is given for the calibration of a high-resolution electronic autocollimator. (paper)

  6. Radiometric modeling and calibration of the Geostationary Imaging Fourier Transform Spectrometer (GIFTS) ground based measurement experiment

    Science.gov (United States)

    Tian, Jialin; Smith, William L.; Gazarik, Michael J.

    2008-12-01

    The ultimate remote sensing benefits of the high resolution Infrared radiance spectrometers will be realized with their geostationary satellite implementation in the form of imaging spectrometers. This will enable dynamic features of the atmosphere's thermodynamic fields and pollutant and greenhouse gas constituents to be observed for revolutionary improvements in weather forecasts and more accurate air quality and climate predictions. As an important step toward realizing this application objective, the Geostationary Imaging Fourier Transform Spectrometer (GIFTS) Engineering Demonstration Unit (EDU) was successfully developed under the NASA New Millennium Program, 2000-2006. The GIFTS-EDU instrument employs three focal plane arrays (FPAs), which gather measurements across the long-wave IR (LWIR), short/mid-wave IR (SMWIR), and visible spectral bands. The GIFTS calibration is achieved using internal blackbody calibration references at ambient (260 K) and hot (286 K) temperatures. In this paper, we introduce a refined calibration technique that utilizes Principle Component (PC) analysis to compensate for instrument distortions and artifacts, therefore, enhancing the absolute calibration accuracy. This method is applied to data collected during the GIFTS Ground Based Measurement (GBM) experiment, together with simultaneous observations by the accurately calibrated AERI (Atmospheric Emitted Radiance Interferometer), both simultaneously zenith viewing the sky through the same external scene mirror at ten-minute intervals throughout a cloudless day at Logan Utah on September 13, 2006. The accurately calibrated GIFTS radiances are produced using the first four PC scores in the GIFTS-AERI regression model. Temperature and moisture profiles retrieved from the PC-calibrated GIFTS radiances are verified against radiosonde measurements collected throughout the GIFTS sky measurement period. Using the GIFTS GBM calibration model, we compute the calibrated radiances from data

  7. 40 CFR 60.1810 - How do I monitor the load of my municipal waste combustion unit?

    Science.gov (United States)

    2010-07-01

    ... municipal waste combustion unit? 60.1810 Section 60.1810 Protection of Environment ENVIRONMENTAL PROTECTION... Guidelines and Compliance Times for Small Municipal Waste Combustion Units Constructed on or Before August 30... combustion unit? (a) If your municipal waste combustion unit generates steam, you must install, calibrate...

  8. Calibration issues for neutron diagnostics

    International Nuclear Information System (INIS)

    Sadler, G.J.; Adams, J.M.; Barnes, C.W.

    1997-01-01

    The performance of diagnostic systems are limited by their weakest constituents, including their calibration issues. Neutron diagnostics are notorious for problems encountered while determining their absolute calibrations, due mainly to the nature of the neutron transport problem. In order to facilitate the determination of an accurate and precise calibration, the diagnostic design should be such as to minimize the scattered neutron flux. ITER will use a comprehensive set of neutron diagnostics--comprising radial and vertical neutron cameras, neutron spectrometers, a neutron activation system and internal and external fission chambers--to provide accurate measurements of fusion power and power densities as a function of time. The calibration of such an important diagnostic system merits careful consideration. Some thoughts have already been given to this subject during the conceptual design phase in relation to the time-integrated neutron activation and time-dependent neutron yield monitors. However, no overall calibration strategy has been worked out so far. This paper represents a first attempt to address this vital issue. Experience gained from present large tokamaks (JET, TFTR and JT60U) and proposals for ITER are reviewed. The need to use a 14-MeV neutron generator as opposed to radioactive sources for in-situ calibration of D-T diagnostics will be stressed. It is clear that the overall absolute determination of fusion power will have to rely on a combination of nuclear measuring techniques, for which the provision of accurate and independent calibrations will constitute an ongoing process as ITER moves from one phase of operation to the next

  9. Improvement of the calibration technique of clinical dosemeters

    International Nuclear Information System (INIS)

    Ehlin Caldas, L.V.

    1988-08-01

    Clinical dosemeters constituted of ionization chambers connected to electrometers are usually calibrated as whole systems in appropriate radiation fields against secondary standard dosemeters in calibration laboratories. This work reports on a technique of component calibration procedures separately for chambers and electrometers applied in the calibration laboratory of IPEN-CNEN, Brazil. For electrometer calibration, redundancy was established by using a standard capacitor of 1000pF (General Radio, USA) and a standard current source based on air ionization with Sr 90 (PTW, Germany). The results from both methods applied to several electrometers of clinical dosemeters agreed within 0.4%. The calibration factors for the respective chambers were determined by intercomparing their response to the response of a certified calibrated chamber in a Co 60 calibration beam using a Keithley electrometer type 617. Overall calibration factors compared with the product of the respective component calibration factors for the tested dosemeters showed an agreement better than 0.7%. This deviation has to be considered with regard to an uncertainty of 2.5% in routine calibration of clinical dosemeters. Calibration by components permits to calibrate ionization chambers one at a time for those hospitals who have several ionization chambers but only one electrometer (small hospitals, hospitals in developing countries). 6 refs, 2 figs, 2 tabs

  10. Calibration curve for germanium spectrometers from solutions calibrated by liquid scintillation counting

    International Nuclear Information System (INIS)

    Grau, A.; Navarro, N.; Rodriguez, L.; Alvarez, A.; Salvador, S.; Diaz, C.

    1996-01-01

    The beta-gamma emitters ''60Co, ''137 Cs, ''131 I, ''210 Pb y ''129 Iare radionuclides for which the calibration by the CIEMAT/NIST method ispossible with uncertainties less than 1%. We prepared, from standardized solutions of these radionuclides, samples in vials of 20 ml. We obtained the calibration curves, efficiency as a function of energy, for two germanium detectors. (Author) 5 refs

  11. Absolute calibration in vivo measurement systems

    International Nuclear Information System (INIS)

    Kruchten, D.A.; Hickman, D.P.

    1991-02-01

    Lawrence Livermore National Laboratory (LLNL) is currently investigating a new method for obtaining absolute calibration factors for radiation measurement systems used to measure internally deposited radionuclides in vivo. Absolute calibration of in vivo measurement systems will eliminate the need to generate a series of human surrogate structures (i.e., phantoms) for calibrating in vivo measurement systems. The absolute calibration of in vivo measurement systems utilizes magnetic resonance imaging (MRI) to define physiological structure, size, and composition. The MRI image provides a digitized representation of the physiological structure, which allows for any mathematical distribution of radionuclides within the body. Using Monte Carlo transport codes, the emission spectrum from the body is predicted. The in vivo measurement equipment is calibrated using the Monte Carlo code and adjusting for the intrinsic properties of the detection system. The calibration factors are verified using measurements of existing phantoms and previously obtained measurements of human volunteers. 8 refs

  12. Calibration of the NPL secondary standard radionuclide calibrator for the new 10R Schott, Type 1+ vials

    International Nuclear Information System (INIS)

    Baker, M.

    2005-01-01

    For many years, P6 vials have been used for the distribution of a wide range of diagnostic and therapeutic radioactive solutions. The activity measurements were performed in this geometry and, in time, the UK calibration system for nuclear medicine was based on this container as a standard. However, one major supplier of radiopharmaceuticals has replaced the P6 vial with the 10R Type 1+ Schott vial. As the dimensions of the new vial are different from those of the P6 vial and the responses of radionuclide calibrators are known to be container dependent, the need for re-calibration became apparent. Preliminary measurements made on some typical radionuclide calibrators for 125 I solution indicated a difference in response of about 10% between the two vials. The master ionisation chamber of the NPL secondary standard radionuclide calibrator has been re-calibrated and new calibration factors and volume correction factors for 10R Schott vials have been derived for the relevant medical radionuclides. The standard holder was also modified to accommodate the new larger vial. The complete list of factors and the method used to determine them is presented in this paper. The availability of these new factors will improve the quality of activity measurements in nuclear medicine, as calibration services can now be provided by NPL for the new container. These factors can also be employed for all commercial NPL secondary standard radionuclide calibrators (now known as the NPL-CRC and previously as the 671 or ISOCAL IV)

  13. Bases for calibration of whole body counters using anthropomorphic physical simulators

    International Nuclear Information System (INIS)

    Dantas, Bernardo Maranhao

    1998-01-01

    The quantification of radionuclides in the human body can be carried out through in vivo measurements performed in facilities generically called whole body counters. The calibration of such units is usually done by using physical anthropomorphic phantoms, which can be defined as artificial structures with geometrical characteristics and attenuation properties similar to the living tissues. This work presents the development of the phantoms necessary to the monitoring of the internal contamination by the radionuclides manipulated in Brazil. It also presents the procedures for the calibration of the detectors used for the in vivo measurements. The developed phantoms are applied in the determination of radionuclides deposited in specific organs, such as Th-232 and Am-241 in the lungs and skull, isotopes of iodine in the thyroid and photon emitters in the energy range from 100 to 3000 keV in the whole body. (author)

  14. Topics in Statistical Calibration

    Science.gov (United States)

    2014-03-27

    Natural cubic spline speed di st 110 B.2 The calibrate function The most basic calibration problem, the one often encountered in more advanced ...0040-1706, 1537-2723. A. M. Mood, F. A. Graybill, and D. C. Boes. Introduction to the Theory of Statistics. McGraw-Hill, Auckland , U.A, 1974. ISBN

  15. Reducing the Variance of Intrinsic Camera Calibration Results in the ROS Camera_Calibration Package

    Science.gov (United States)

    Chiou, Geoffrey Nelson

    The intrinsic calibration of a camera is the process in which the internal optical and geometric characteristics of the camera are determined. If accurate intrinsic parameters of a camera are known, the ray in 3D space that every point in the image lies on can be determined. Pairing with another camera allows for the position of the points in the image to be calculated by intersection of the rays. Accurate intrinsics also allow for the position and orientation of a camera relative to some world coordinate system to be calculated. These two reasons for having accurate intrinsic calibration for a camera are especially important in the field of industrial robotics where 3D cameras are frequently mounted on the ends of manipulators. In the ROS (Robot Operating System) ecosystem, the camera_calibration package is the default standard for intrinsic camera calibration. Several researchers from the Industrial Robotics & Automation division at Southwest Research Institute have noted that this package results in large variances in the intrinsic parameters of the camera when calibrating across multiple attempts. There are also open issues on this matter in their public repository that have not been addressed by the developers. In this thesis, we confirm that the camera_calibration package does indeed return different results across multiple attempts, test out several possible hypothesizes as to why, identify the reason, and provide simple solution to fix the cause of the issue.

  16. Update on Calibration of the Lawrence Livermore National Laboratory Passive-Active Neutron Drum Shuffler for Measurement of Highly Enriched Uranium Oxide

    International Nuclear Information System (INIS)

    Mount, M.; O'Connell, W.; Cochran, C.; Rinard, P.; Dearborn, D.; Endres, E.

    2002-01-01

    In October of 1999, Lawrence Livermore National Laboratory (LLNL) began an effort to calibrate the LLNL passive-active neutron (PAN) drum shuffler for measurement of highly enriched uranium (HEU) oxide. A single unit of certified reference material (CRM) 149 (Uranium (93% Enriched) Oxide - U 3 O 8 Standard for Neutron Counting Measurements) was used to (1) develop a mass calibration curve for HEU oxide in the nominal range of 393 g to 3144 g 235 U, and (2) perform a detailed axial and radial mapping of the detector response over a wide region of the PAN shuffler counting chamber. Results from these efforts were reported at the Institute of Nuclear Materials Management 4lSt Annual Meeting in July 2000. This paper describes subsequent efforts by LLNL to use a unit of CRM 146 (Uranium Isotopic Standard for Gamma Spectrometry Measurements) in consort with Monte Carlo simulations of the PAN shuffler response to CRM 149 and CRM 146 units and a selected set of containers with CRM 149-equivalent U 3 O 8 to (1) extend the low range of the reported mass calibration curve to 10 g 235 U, (2) evaluate the effect of U 3 O 8 density (2.4 g/cm 3 to 4.8 g/cm 3 ) and container size (5.24 cm to 12.17 cm inside diameter and 6.35 cm to 17.72 cm inside height) on the PAN shuffler response, and (3) develop mass calibration curves for U 3 O 8 enriched to 20.1 wt% 235 U and 52.5 wt% 235 U.

  17. Calibrating the neutron moisture meter: Precision and economy

    International Nuclear Information System (INIS)

    Akhter, J.; Waheed, R.A.; Hignett, C.T.; Greacen, E.L.

    2000-01-01

    Established laboratory and field calibration procedures for the neutron moisture meter are demonstrated on a uniform soil and alternative, low cost procedures on a duplex, less uniform soil. The effect of field variability on the calibration methodology is discussed with the aim of optimising calibration reliability at minimal cost. The difference between calibration for a soil material, or for a field (a range of soil materials) is considered. In particular, calibration for the estimation of water content change is shown to be a different problem from calibration for the estimation of water content in a variable field. Techniques aimed at detecting field variability problems during calibration are suggested, and methods for optimising the results for the intended use of the instrument are outlined. Pairing of calibration tubes, alternative methods of analysis of calibration data, and use of other information from the field to measure its variability, can improve the precision of calibration procedures to the point where minimal calibration effort, with careful analysis, can provide reliable results. (author)

  18. Calibrating page sized Gafchromic EBT3 films

    International Nuclear Information System (INIS)

    Crijns, W.; Maes, F.; Heide, U. A. van der; Van den Heuvel, F.

    2013-01-01

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittance values. The linear combination model, combined a monomer transmittance state (T 0 ) and a polymer transmittance state (T ∞ ) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the optimal balance between

  19. Research on the method of establishing the total radiation meter calibration device

    Science.gov (United States)

    Gao, Jianqiang; Xia, Ming; Xia, Junwen; Zhang, Dong

    2015-10-01

    Pyranometer is an instrument used to measure the solar radiation, according to pyranometer differs as installation state, can be respectively measured total solar radiation, reflected radiation, or with the help of shading device for measuring scattering radiation. Pyranometer uses the principle of thermoelectric effect, inductive element adopts winding plating type multi junction thermopile, its surface is coated with black coating with high absorption rate. Hot junction in the induction surface, while the cold junction is located in the body, the cold and hot junction produce thermoelectric potential. In the linear range, the output signal is proportional to the solar irradiance. Traceability to national meteorological station, as the unit of the national legal metrology organizations, the responsibility is to transfer value of the sun and the earth radiation value about the national meteorological industry. Using the method of comparison, with indoor calibration of solar simulator, at the same location, standard pyranometer and measured pyranometer were alternately measured radiation irradiance, depending on the irradiation sensitivity standard pyranometer were calculated the radiation sensitivity of measured pyranometer. This paper is mainly about the design and calibration method of the pyranometer indoor device. The uncertainty of the calibration result is also evaluated.

  20. Compact blackbody calibration sources for in-flight calibration of spaceborne infrared instruments

    Science.gov (United States)

    Scheiding, S.; Driescher, H.; Walter, I.; Hanbuch, K.; Paul, M.; Hartmann, M.; Scheiding, M.

    2017-11-01

    High-emissivity blackbodies are mandatory as calibration sources in infrared radiometers. Besides the requirements on the high spectral emissivity and low reflectance, constraints regarding energy consumption, installation space and mass must be considered during instrument design. Cavity radiators provide an outstanding spectral emissivity to the price of installation space and mass of the calibration source. Surface radiation sources are mainly limited by the spectral emissivity of the functional coating and the homogeneity of the temperature distribution. The effective emissivity of a "black" surface can be optimized, by structuring the substrate with the aim to enlarge the ratio of the surface to its projection. Based on the experiences of the Mercury Radiometer and Thermal Infrared Spectrometer (MERTIS) calibration source MBB3, the results of the surface structuring on the effective emissivity are described analytically and compared to the experimental performance. Different geometries are analyzed and the production methods are discussed. The high-emissivity temperature calibration source features values of 0.99 for wavelength from 5 μm to 10 μm and emissivity larger than 0.95 for the spectral range from 10 μm to 40 μm.

  1. On the use of mobile phones and wearable microphones for noise exposure measurements: Calibration and measurement accuracy

    Science.gov (United States)

    Dumoulin, Romain

    Despite the fact that noise-induced hearing loss remains the number one occupational disease in developed countries, individual noise exposure levels are still rarely known and infrequently tracked. Indeed, efforts to standardize noise exposure levels present disadvantages such as costly instrumentation and difficulties associated with on site implementation. Given their advanced technical capabilities and widespread daily usage, mobile phones could be used to measure noise levels and make noise monitoring more accessible. However, the use of mobile phones for measuring noise exposure is currently limited due to the lack of formal procedures for their calibration and challenges regarding the measurement procedure. Our research investigated the calibration of mobile phone-based solutions for measuring noise exposure using a mobile phone's built-in microphones and wearable external microphones. The proposed calibration approach integrated corrections that took into account microphone placement error. The corrections were of two types: frequency-dependent, using a digital filter and noise level-dependent, based on the difference between the C-weighted noise level minus A-weighted noise level of the noise measured by the phone. The electro-acoustical limitations and measurement calibration procedure of the mobile phone were investigated. The study also sought to quantify the effect of noise exposure characteristics on the accuracy of calibrated mobile phone measurements. Measurements were carried out in reverberant and semi-anechoic chambers with several mobiles phone units of the same model, two types of external devices (an earpiece and a headset with an in-line microphone) and an acoustical test fixture (ATF). The proposed calibration approach significantly improved the accuracy of the noise level measurements in diffuse and free fields, with better results in the diffuse field and with ATF positions causing little or no acoustic shadowing. Several sources of errors

  2. New thermal neutron calibration channel at LNMRI/IRD

    Energy Technology Data Exchange (ETDEWEB)

    Astuto, A.; Lopes, R.T., E-mail: achillesbr@gmail.com [Coordenacao dos Programas de Pos-Graduacao em Engenharia (COPPE/UFRJ), Rio de Janeiro, RJ (Brazil); Patrao, K.C.S.; Fonseca, E.S.; Pereira, W.W. [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ/LNMRI), Rio de Janeiro, RJ (Brazil). Lab. Nacional de Metrologia das Radiacoes Ionizantes

    2015-07-01

    A new standard thermal neutron flux unit was designed in the National Ionizing Radiation Metrology Laboratory (LNMRI) for calibration of neutron detectors. Fluence is achieved by moderation of four {sup 241}Am-Be sources with 596 GBq each, in a facility built with graphite and paraffin blocks. The study was divided into two stages. First, simulations were performed using MCNPX code in different geometric arrangements, seeking the best performance in terms of fluence and their uncertainties. Last, the system was assembled based on the results obtained on the simulations. The simulation results indicate quasi-homogeneous fluence (less than 1%) in the central chamber. (author)

  3. CryoSat SIRAL Calibration and Performance

    Science.gov (United States)

    Fornari, Marco; Scagliola, Michele; Tagliani, Nicolas; Parrinello, Tommaso

    2013-04-01

    The main payload of CryoSat is a Ku band pulse-width limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for azimuth processing. This allows to reach an along track resolution of about 250 meters which is a significant improvement over traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed, including both an internal and external calibration. The internal calibration monitors the instrument impulse response and the transfer function, like traditional altimeters. In addition to that, the interferometer requires a special calibration developed ad hoc for SIRAL. The external calibration is performed with the use of a ground transponder, located in Svalbard, which receives SIRAL signal and sends the echo back to the satellite. Internal calibration data are processed on ground by the CryoSat Instrument Processing Facility (IPF1) and then applied to the science data. By April 2013, almost 3 years of calibration data will be available, which will be shown in this poster. The external calibration (transponder) data are processed and analyzed independently from the operational chain. The use of an external transponder has been very useful to determine instrument performance and for the tuning of the on-ground processor. This poster presents the transponder results in terms of range noise and datation error.

  4. Features calibration of the dynamic force transducers

    Science.gov (United States)

    Sc., M. Yu Prilepko D.; Lysenko, V. G.

    2018-04-01

    The article discusses calibration methods of dynamic forces measuring instruments. The relevance of work is dictated by need to valid definition of the dynamic forces transducers metrological characteristics taking into account their intended application. The aim of this work is choice justification of calibration method, which provides the definition dynamic forces transducers metrological characteristics under simulation operating conditions for determining suitability for using in accordance with its purpose. The following tasks are solved: the mathematical model and the main measurements equation of calibration dynamic forces transducers by load weight, the main budget uncertainty components of calibration are defined. The new method of dynamic forces transducers calibration with use the reference converter “force-deformation” based on the calibrated elastic element and measurement of his deformation by a laser interferometer is offered. The mathematical model and the main measurements equation of the offered method is constructed. It is shown that use of calibration method based on measurements by the laser interferometer of calibrated elastic element deformations allows to exclude or to considerably reduce the uncertainty budget components inherent to method of load weight.

  5. Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Faber, M.H.; Sørensen, John Dalsgaard

    2003-01-01

    The present paper addresses fundamental concepts of reliability based code calibration. First basic principles of structural reliability theory are introduced and it is shown how the results of FORM based reliability analysis may be related to partial safety factors and characteristic values....... Thereafter the code calibration problem is presented in its principal decision theoretical form and it is discussed how acceptable levels of failure probability (or target reliabilities) may be established. Furthermore suggested values for acceptable annual failure probabilities are given for ultimate...... and serviceability limit states. Finally the paper describes the Joint Committee on Structural Safety (JCSS) recommended procedure - CodeCal - for the practical implementation of reliability based code calibration of LRFD based design codes....

  6. Calibration Lessons Learned from Hyperion Experience

    Science.gov (United States)

    Casement, S.; Ho, K.; Sandor-Leahy, S.; Biggar, S.; Czapla-Myers, J.; McCorkel, J.; Thome, K.

    2009-12-01

    The use of hyperspectral imagers to provide climate-quality data sets, such as those expected from the solar reflective sensor on the Climate Absolute Radiance and Refractivity Observatory (CLARREO), requires stringent radiometric calibration requirements. These stringent requirements have been nearly met with broadband radiometers such as CERES, but high resolution spectrometers pose additional challenges. A review of the calibration processes for past space-based HSIs provide guidance on the calibration processes that will be needed for future sensors. In November 2000, the Earth Observer-1 (EO-1) platform was launched onboard a Boeing Delta II launch vehicle. The primary purpose of the EO-1 mission was to provide a technological testbed for spaceborne components. The platform has three sensors onboard, of which, the hyperspectral imager (HSI) Hyperion, is discussed here. The Hyperion sensor at the time had no comparable sensor in earth orbit, being the first grating-based, hyperspectral, civilian sensor in earth orbit. Ground and on-orbit calibration procedures including all cross-calibration activities have achieved an estimated instrument absolute radiometric error of 2.9% in the Visible channel (0.4 - 1.0 microns) and 3.4% in the shortwave infrared (SWIR, 0.9 - 2.5 microns) channel (EO-1/Hyperion Early Orbit Checkout Report Part II On-Orbit Performance Verification and Calibration). This paper describes the key components of the Hyperion calibration process that are applicable to future HSI missions. The pre-launch methods relied on then newly-developed, detector-based methods. Subsequent vicarious methods including cross-calibration with other sensors and the reflectance-based method showed significant differences from the prelaunch calibration. Such a difference demonstrated the importance of the vicarious methods as well as pointing to areas for improvement in the prelaunch methods. We also identify areas where lessons learned from Hyperion regarding

  7. DEM Calibration Approach: design of experiment

    Science.gov (United States)

    Boikov, A. V.; Savelev, R. V.; Payor, V. A.

    2018-05-01

    The problem of DEM models calibration is considered in the article. It is proposed to divide models input parameters into those that require iterative calibration and those that are recommended to measure directly. A new method for model calibration based on the design of the experiment for iteratively calibrated parameters is proposed. The experiment is conducted using a specially designed stand. The results are processed with technical vision algorithms. Approximating functions are obtained and the error of the implemented software and hardware complex is estimated. The prospects of the obtained results are discussed.

  8. Exploitation of secondary standard for calibration in units of Dw,1cm and assessment of several HDR brachytherapy planning systems

    International Nuclear Information System (INIS)

    Gabris, F.; Zeman, J.; Valenta, J.; Gabris, F.; Selbach, H.J.

    2012-01-01

    A secondary standard of the BEV, calibrated at the PTB in terms of D w,1c m, was used for calibration of the well-type chamber-based measuring systems used in clinics. In addition to the calibration, we tried to employ it for assessment of treatment planning systems (TPS) used for each particular after-loader. The dose to water at 1 cm distance from the source position was calculated by the TPS, using reference data from the source producer certificate. The values were compared directly with the dose measured at the same distance from the source. The comparison has been carried out for GammaMed Plus and MicroSelectron HDR sources. Differences of secondary standard measurements and TPS calculations were lower than ±5%, which is below the achievable uncertainty of both dose measurement and dose determination by the TPS. Nevertheless, it is higher than generally accepted in the case of external beam radiotherapy. Additional direct measurements in terms of D w,1c m may improve the safety and reliability of patient treatment. (authors)

  9. In-situ calibration of clinical built-in KAP meters with traceability to a primary standard using a reference KAP meter.

    Science.gov (United States)

    Malusek, A; Helmrot, E; Sandborg, M; Grindborg, J-E; Carlsson, G Alm

    2014-12-07

    The air kerma-area product (KAP) is used for settings of diagnostic reference levels. The International Atomic Energy Agency (IAEA) recommends that doses in diagnostic radiology (including the KAP values) be estimated with an accuracy of at least ± 7% (k = 2). Industry standards defined by the International Electrotechnical Commission (IEC) specify that the uncertainty of KAP meter measurements should be less than ± 25% (k = 2). Medical physicists willing to comply with the IAEA's recommendation need to apply correction factors to KAP values reported by x-ray units. The aim of this work is to present and evaluate a calibration method for built-in KAP meters on clinical x-ray units. The method is based on (i) a tandem calibration method, which uses a reference KAP meter calibrated to measure the incident radiation, (ii) measurements using an energy-independent ionization chamber to correct for the energy dependence of the reference KAP meter, and (iii) Monte Carlo simulations of the beam quality correction factors that correct for differences between beam qualities at a standard laboratory and the clinic. The method was applied to the KAP meter in a Siemens Aristos FX plus unit. It was found that values reported by the built-in KAP meter differed from the more accurate values measured by the reference KAP meter by more than 25% for high tube voltages (more than 140 kV) and heavily filtered beams (0.3 mm Cu). Associated uncertainties were too high to claim that the IEC's limit of 25% was exceeded. Nevertheless the differences were high enough to justify the need for a more accurate calibration of built-in KAP meters.

  10. Calibration of moisture monitors

    International Nuclear Information System (INIS)

    Gutierrez, R.L.

    1979-02-01

    A method for calibrating an aluminum oxide hygrometer against an optical chilled mirror dew-point hygrometer has been established. A theoretical cross-point line of dew points from both hygrometers and a maximum moisture content of 10 ppM/sub v/ are used to define an area for calibrating the sensor probes of the aluminum oxide hygrometer

  11. Nuclear radiation moisture gauge calibration standard

    International Nuclear Information System (INIS)

    1977-01-01

    A hydrophobic standard for calibrating nuclear radiation moisture gauges is described. Each standard has physical characteristics and dimensions effective for representing to a nuclear gauge undergoing calibration, an infinite mass of homogeneous hydrogen content. Calibration standards are discussed which are suitable for use with surface gauges and with depth gauges. (C.F.)

  12. MAVEN SEP Calibrated Data Bundle

    Data.gov (United States)

    National Aeronautics and Space Administration — The maven.sep.calibrated Level 2 Science Data Bundle contains fully calibrated SEP data, as well as the raw count data from which they are derived, and ancillary...

  13. Development of a generic auto-calibration package for regional ecological modeling and application in the Central Plains of the United States

    Science.gov (United States)

    Wu, Yiping; Liu, Shuguang; Li, Zhengpeng; Dahal, Devendra; Young, Claudia J.; Schmidt, Gail L.; Liu, Jinxun; Davis, Brian; Sohl, Terry L.; Werner, Jeremy M.; Oeding, Jennifer

    2014-01-01

    Process-oriented ecological models are frequently used for predicting potential impacts of global changes such as climate and land-cover changes, which can be useful for policy making. It is critical but challenging to automatically derive optimal parameter values at different scales, especially at regional scale, and validate the model performance. In this study, we developed an automatic calibration (auto-calibration) function for a well-established biogeochemical model—the General Ensemble Biogeochemical Modeling System (GEMS)-Erosion Deposition Carbon Model (EDCM)—using data assimilation technique: the Shuffled Complex Evolution algorithm and a model-inversion R package—Flexible Modeling Environment (FME). The new functionality can support multi-parameter and multi-objective auto-calibration of EDCM at the both pixel and regional levels. We also developed a post-processing procedure for GEMS to provide options to save the pixel-based or aggregated county-land cover specific parameter values for subsequent simulations. In our case study, we successfully applied the updated model (EDCM-Auto) for a single crop pixel with a corn–wheat rotation and a large ecological region (Level II)—Central USA Plains. The evaluation results indicate that EDCM-Auto is applicable at multiple scales and is capable to handle land cover changes (e.g., crop rotations). The model also performs well in capturing the spatial pattern of grain yield production for crops and net primary production (NPP) for other ecosystems across the region, which is a good example for implementing calibration and validation of ecological models with readily available survey data (grain yield) and remote sensing data (NPP) at regional and national levels. The developed platform for auto-calibration can be readily expanded to incorporate other model inversion algorithms and potential R packages, and also be applied to other ecological models.

  14. Spectrophotometric calibration system for DECam

    Science.gov (United States)

    Rheault, J.-P.; DePoy, D. L.; Marshall, J. L.; Prochaska, T.; Allen, R.; Wise, J.; Martin, E.; Williams, P.

    2012-09-01

    We describe a spectrophotometric calibration system that is being implemented as part of the DES DECam project at the Blanco 4 meter at CTIO. Our calibration system uses a 1nm wide tunable source to measure the instrumental response function of the telescope optics and detector from 300nm up to 1100nm. This calibration will be performed regularly to monitor any change in the transmission function of the telescope during the 5 year survey. The system consists of a monochromator based tunable light source that provides illumination on a dome flat that is monitored by calibrated photodiodes that allow us to measure the telescope throughput as a function of wavelength. Our system has a peak output power of 2 mW, equivalent to a flux of approximately 800 photons/s/pixel on DECam.

  15. Calibration of Yucca Mountain unsaturated zone flow and transport model using porewater chloride data

    International Nuclear Information System (INIS)

    Liu, Jianchun; Sonnenthal, Eric L.; Bodvarsson, Gudmundur S.

    2002-01-01

    In this study, porewater chloride data from Yucca Mountain, Nevada, are analyzed and modeled by 3-D chemical transport simulations and analytical methods. The simulation modeling approach is based on a continuum formulation of coupled multiphase fluid flow and tracer transport processes through fractured porous rock, using a dual-continuum concept. Infiltration-rate calibrations were using the pore water chloride data. Model results of chloride distributions were improved in matching the observed data with the calibrated infiltration rates. Statistical analyses of the frequency distribution for overall percolation fluxes and chloride concentration in the unsaturated zone system demonstrate that the use of the calibrated infiltration rates had insignificant effect on the distribution of simulated percolation fluxes but significantly changed the predicated distribution of simulated chloride concentrations. An analytical method was also applied to model transient chloride transport. The method was verified by 3-D simulation results as able to capture major chemical transient behavior and trends. Effects of lateral flow in the Paintbrush nonwelded unit on percolation fluxes and chloride distribution were studied by 3-D simulations with increased horizontal permeability. The combined results from these model calibrations furnish important information for the UZ model studies, contributing to performance assessment of the potential repository

  16. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 2: Data standardization for tank calibration

    International Nuclear Information System (INIS)

    2007-01-01

    Measurements of the volume and height of liquid in a process accountancy tank are often made in order to estimate or verify the tank's calibration or volume measurement equation. The calibration equation relates the response of the tank's measurement system to some independent measure of tank volume. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation (the inverse of the calibration equation), which relates tank volume to measurement system response. In this part of ISO 18213, it is assumed that the primary measurement-system response variable is liquid height and that the primary measure of liquid content is volume. This part of ISO 18213 presents procedures for standardizing a set of calibration data to a fixed set of reference conditions so as to minimize the effect of variations in ambient conditions that occur during the measurement process. The procedures presented herein apply generally to measurements of liquid height and volume obtained for the purpose of calibrating a tank (i.e. calibrating a tank's measurement system). When used in connection with other parts of ISO 18213, these procedures apply specifically to tanks equipped with bubbler probe systems for measuring liquid content. The standardization algorithms presented herein can be profitably applied when only estimates of ambient conditions, such as temperature, are available. However, the most reliable results are obtained when relevant ambient conditions are measured for each measurement of volume and liquid height in a set of calibration data. Information is provided on scope, physical principles, data required, calibration data, dimensional changes in the tank, multiple calibration runs and results on standardized calibration data. Four annexes inform about density of water, buoyancy corrections for mass determination, determination of tank heel volume and statistical method for aligning data from several calibration runs. A bibliography is

  17. Calibration biases in logical reasoning tasks

    Directory of Open Access Journals (Sweden)

    Guillermo Macbeth

    2013-08-01

    Full Text Available The aim of this contribution is to present an experimental study about calibration in deductive reasoning tasks. Calibration is defi ned as the empirical convergence or divergence between the objective and the subjective success. The underconfi dence bias is understood as the dominance of the former over the latter. The hypothesis of this study states that the form of the propositions presented in the experiment is critical for calibration phenomena. Affi rmative and negative propositions are distinguished in their cognitive processing. Results suggests that monotonous compound propositions are prone to underconfi dence. An heuristic approach to this phenomenon is proposed. The activation of a monotony heuristic would produce an illusion of simplicity that generates the calibration bias. These evidence is analysed in the context of the metacognitive modeling of calibration phenomena.

  18. Calibrating page sized Gafchromic EBT3 films

    Energy Technology Data Exchange (ETDEWEB)

    Crijns, W.; Maes, F.; Heide, U. A. van der; Van den Heuvel, F. [Department of Radiation Oncology, University Hospitals Leuven, Herestraat 49, 3000 Leuven (Belgium); Department ESAT/PSI-Medical Image Computing, Medical Imaging Research Center, KU Leuven, Herestraat 49, 3000 Leuven (Belgium); Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Plesmanlaan 121, 1066 CX Amsterdam (Netherlands); Department of Radiation Oncology, University Hospitals Leuven, Herestraat 49, 3000 Leuven (Belgium)

    2013-01-15

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittance values. The linear combination model, combined a monomer transmittance state (T{sub 0}) and a polymer transmittance state (T{sub {infinity}}) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the optimal

  19. Calibration of ionization chamber survey meter

    International Nuclear Information System (INIS)

    Kadhim, A.K.; Kadni, T.B.

    2016-01-01

    Radiation measuring devices need to process calibration which lose their sensitivity and the extent of the response and the amount of stability under a changing conditions from time to time and this period depends on the nature and use of field in which used devices. A comparison study was done toa (45 I P) ( ionization chamber survey meter) and this showed the variation factor in five different years. This study also displayed the concept of radiation instrument calibration and necessity of every year calibration of them.In this project we used the five years calibration data for ionization chamber survey meter model Inspector (1/C F). the value deviation (∆ %) of Cfs for four years of calibration in comparison of C F for the year 2007 are very high and the device under research is not good to use in field and reliable because the ionization chamber is very sensitive to humidity and must calibrate a year or less, or due ing every two years and must maintain carefully to reduce the discarded effects the measurements.

  20. Calibration Of Partial-Pressure-Of-Oxygen Sensors

    Science.gov (United States)

    Yount, David W.; Heronimus, Kevin

    1995-01-01

    Report and analysis of, and discussion of improvements in, procedure for calibrating partial-pressure-of-oxygen sensors to satisfy Spacelab calibration requirements released. Sensors exhibit fast drift, which results in short calibration period not suitable for Spacelab. By assessing complete process of determining total drift range available, calibration procedure modified to eliminate errors and still satisfy requirements without compromising integrity of system.

  1. Chaos, Consternation and CALIPSO Calibration: New Strategies for Calibrating the CALIOP 1064 nm Channel

    Science.gov (United States)

    Vaughan, Mark; Garnier, Anne; Liu, Zhaoyan; Josset, Damien; Hu, Yongxiang; Lee, Kam-Pui; Hunt, William; Vernier, Jean-Paul; Rodier, Sharon; Pelon, Jaques; hide

    2012-01-01

    The very low signal-to-noise ratios of the 1064 nm CALIOP molecular backscatter signal make it effectively impossible to employ the "clear air" normalization technique typically used to calibrate elastic back-scatter lidars. The CALIPSO mission has thus chosen to cross-calibrate their 1064 nm measurements with respect to the 532 nm data using the two-wavelength backscatter from cirrus clouds. In this paper we discuss several known issues in the version 3 CALIOP 1064 nm calibration procedure, and describe the strategies that will be employed in the version 4 data release to surmount these problems.

  2. Standardization of calibration method of whole-body counter. 1. Calibration by using anthropometric phantoms

    International Nuclear Information System (INIS)

    Ishikawa, Tetsuo; Matsumoto, Masaki; Uchiyama, Masafumi; Kobayashi, Sadayoshi; Mizushita, Seiichi.

    1995-01-01

    To standardize the calibration methods of whole-body counters, three anthropometric phantoms were manufactured based on dozens of Japanese average value of body size data. Using these phantoms, the calibrations of some whole-body counters were carried out and the comparison of counting efficiency between anthropometric phantoms and block phantoms, which used to be used for the calibration of whole-body counters generally, was implemented. Five whole-body counters, one scanning system, two stationary systems and two chair systems, were used for this study. The following results were derived: As an example, in NIRS scanning system, the counting efficiency of anthropometric phantom of 162cm height was 12.7% greater than that of block phantom of the same height. This means 137 Cs body burdens in adult men used to be estimated with the excess of about 10%. Body burdens tended to be estimated excessively in adult because the difference of counting efficiency between anthropometric phantom and block phantom increases with increase of height. To standardize body burden data measured with various whole-body counters, the calibration of each whole-body counter should be conducted using anthropometric phantoms and phantoms which used to be used for the calibration of that whole-body counter. (author)

  3. Vicarious Calibration of Beijing-1 Multispectral Imagers

    Directory of Open Access Journals (Sweden)

    Zhengchao Chen

    2014-02-01

    Full Text Available For on-orbit calibration of the Beijing-1 multispectral imagers (Beijing-1/MS, a field calibration campaign was performed at the Dunhuang calibration site during September and October of 2008. Based on the in situ data and images from Beijing-1 and Terra/Moderate Resolution Imaging Spectroradiometer (MODIS, three vicarious calibration methods (i.e., reflectance-based, irradiance-based, and cross-calibration were used to calculate the top-of-atmosphere (TOA radiance of Beijing-1. An analysis was then performed to determine or identify systematic and accidental errors, and the overall uncertainty was assessed for each individual method. The findings show that the reflectance-based method has an uncertainty of more than 10% if the aerosol optical depth (AOD exceeds 0.2. The cross-calibration method is able to reach an error level within 7% if the images are selected carefully. The final calibration coefficients were derived from the irradiance-based data for 6 September 2008, with an uncertainty estimated to be less than 5%.

  4. Calibration and simulation of Heston model

    Directory of Open Access Journals (Sweden)

    Mrázek Milan

    2017-05-01

    Full Text Available We calibrate Heston stochastic volatility model to real market data using several optimization techniques. We compare both global and local optimizers for different weights showing remarkable differences even for data (DAX options from two consecutive days. We provide a novel calibration procedure that incorporates the usage of approximation formula and outperforms significantly other existing calibration methods.

  5. Evaluation of Hydrologic Simulations Developed Using Multi-Model Synthesis and Remotely-Sensed Data within a Portfolio of Calibration Strategies

    Science.gov (United States)

    Lafontaine, J.; Hay, L.; Markstrom, S. L.

    2016-12-01

    The United States Geological Survey (USGS) has developed a National Hydrologic Model (NHM) to support coordinated, comprehensive and consistent hydrologic model development, and facilitate the application of hydrologic simulations within the conterminous United States (CONUS). As many stream reaches in the CONUS are either not gaged, or are substantially impacted by water use or flow regulation, ancillary information must be used to determine reasonable parameter estimations for streamflow simulations. Hydrologic models for 1,576 gaged watersheds across the CONUS were developed to test the feasibility of improving streamflow simulations linking physically-based hydrologic models with remotely-sensed data products (i.e. snow water equivalent). Initially, the physically-based models were calibrated to measured streamflow data to provide a baseline for comparison across multiple calibration strategy tests. In addition, not all ancillary datasets are appropriate for application to all parts of the CONUS (e.g. snow water equivalent in the southeastern U.S., where snow is a rarity). As it is not expected that any one data product or model simulation will be sufficient for representing hydrologic behavior across the entire CONUS, a systematic evaluation of which data products improve hydrologic simulations for various regions across the CONUS was performed. The resulting portfolio of calibration strategies can be used to guide selection of an appropriate combination of modeled and measured information for hydrologic model development and calibration. In addition, these calibration strategies have been developed to be flexible so that new data products can be assimilated. This analysis provides a foundation to understand how well models work when sufficient streamflow data are not available and could be used to further inform hydrologic model parameter development for ungaged areas.

  6. Compact Optical Technique for Streak Camera Calibration

    International Nuclear Information System (INIS)

    Bell, P; Griffith, R; Hagans, K; Lerche, R; Allen, C; Davies, T; Janson, F; Justin, R; Marshall, B; Sweningsen, O

    2004-01-01

    to a streak camera just before and at shot time to verify that the streak camera sweep unit is functioning properly. A comb signal applied at shot time does not have the statistical quality of the multiple recordings used to calibrate a streak camera time base. However, it does allow the user to verify that the correct sweep card is installed and that it is functioning within about 1.5% of its expected speed2. The principle component in the comb generator is a custom packaged Vertical Cavity Surface Emitting Laser (VCSEL), Figure 2. The laser is modulated either directly by a crystal controlled sine wave oscillator or by an impulse generator that is supplied by the oscillator, Figure 3. A pulse bias is added to the sine wave or impulse input to bring the VCSEL to lasing for the time interval that the optical comb train is required. The pulse bias can be used to adjust the intensity of the optical pulse train. An example of a 3 GHz comb streak is shown in Figure 4(a)

  7. Infrared stereo calibration for unmanned ground vehicle navigation

    Science.gov (United States)

    Harguess, Josh; Strange, Shawn

    2014-06-01

    The problem of calibrating two color cameras as a stereo pair has been heavily researched and many off-the-shelf software packages, such as Robot Operating System and OpenCV, include calibration routines that work in most cases. However, the problem of calibrating two infrared (IR) cameras for the purposes of sensor fusion and point could generation is relatively new and many challenges exist. We present a comparison of color camera and IR camera stereo calibration using data from an unmanned ground vehicle. There are two main challenges in IR stereo calibration; the calibration board (material, design, etc.) and the accuracy of calibration pattern detection. We present our analysis of these challenges along with our IR stereo calibration methodology. Finally, we present our results both visually and analytically with computed reprojection errors.

  8. CR-39 nuclear track detector used for neutron dosimetry: system calibration

    International Nuclear Information System (INIS)

    Saint Martin, G.; Lopez, F.; Bernaola, Omar A.

    2009-01-01

    Stacks composed by 1 mm thickness CR-39 foils and polyethylene and PVC films were evaluated to be used as neutron dosemeters. Irradiations were made with a calibrated 241 Am-Be source in a dose range from 0 to 3.1 mSv and the etching conditions were optimized. The measurements of number of tracks per surface unit in the CR-39 detectors showed a good linear behaviour as a function of the dose. The minimum detectable dose equivalent (MDDE) was calculated. (author)

  9. Vessel calibration for accurate material accountancy at RRP

    International Nuclear Information System (INIS)

    Yanagisawa, Yuu; Ono, Sawako; Iwamoto, Tomonori

    2004-01-01

    RRP has a 800t·Upr capacity a year to re-process, where would be handled a large amount of nuclear materials as solution. A large scale plant like RRP will require accurate materials accountancy system, so that the vessel calibration with high-precision is very important as initial vessel calibration before operation. In order to obtain the calibration curve, it is needed well-known each the increment volume related with liquid height. Then we performed at least 2 or 3 times run with water for vessel calibration and careful evaluation for the calibration data should be needed. We performed vessel calibration overall 210 vessels, and the calibration of 81 vessels including IAT and OAT were held under presence of JSGO and IAEA inspectors taking into account importance on the material accountancy. This paper describes outline of the initial vessel calibration and calibration results based on back pressure measurement with dip tubes. (author)

  10. COMPARISON OF METHODS FOR GEOMETRIC CAMERA CALIBRATION

    Directory of Open Access Journals (Sweden)

    J. Hieronymus

    2012-09-01

    Full Text Available Methods for geometric calibration of cameras in close-range photogrammetry are established and well investigated. The most common one is based on test-fields with well-known pattern, which are observed from different directions. The parameters of a distortion model are calculated using bundle-block-adjustment-algorithms. This methods works well for short focal lengths, but is essentially more problematic to use with large focal lengths. Those would require very large test-fields and surrounding space. To overcome this problem, there is another common method for calibration used in remote sensing. It employs measurements using collimator and a goniometer. A third calibration method uses diffractive optical elements (DOE to project holograms of well known pattern. In this paper these three calibration methods are compared empirically, especially in terms of accuracy. A camera has been calibrated with those methods mentioned above. All methods provide a set of distortion correction parameters as used by the photogrammetric software Australis. The resulting parameter values are very similar for all investigated methods. The three sets of distortion parameters are crosscompared against all three calibration methods. This is achieved by inserting the gained distortion parameters as fixed input into the calibration algorithms and only adjusting the exterior orientation. The RMS (root mean square of the remaining image coordinate residuals are taken as a measure of distortion correction quality. There are differences resulting from the different calibration methods. Nevertheless the measure is small for every comparison, which means that all three calibration methods can be used for accurate geometric calibration.

  11. Pre-flight calibration and initial data processing for the ChemCam laser-induced breakdown spectroscopy instrument on the Mars Science Laboratory rover

    Energy Technology Data Exchange (ETDEWEB)

    Wiens, R.C., E-mail: rwiens@lanl.gov [Los Alamos National Laboratory, Los Alamos, NM 87544 (United States); Maurice, S.; Lasue, J.; Forni, O. [Institut de Recherche en Astrophysique et Planetologie, Toulouse (France); Anderson, R.B. [United States Geological Survey, Flagstaff, AZ (United States); Clegg, S. [Los Alamos National Laboratory, Los Alamos, NM 87544 (United States); Bender, S. [Planetary Science Institute, Tucson, AZ (United States); Blaney, D. [Jet Propulsion Laboratory, Pasadena, CA (United States); Barraclough, B.L. [Planetary Science Institute, Tucson, AZ (United States); Cousin, A. [Los Alamos National Laboratory, Los Alamos, NM 87544 (United States); Institut de Recherche en Astrophysique et Planetologie, Toulouse (France); Deflores, L. [Jet Propulsion Laboratory, Pasadena, CA (United States); Delapp, D. [Los Alamos National Laboratory, Los Alamos, NM 87544 (United States); Dyar, M.D. [Mount Holyoke College, South Hadley, MA (United States); Fabre, C. [Georessources, Nancy (France); Gasnault, O. [Institut de Recherche en Astrophysique et Planetologie, Toulouse (France); Lanza, N. [Los Alamos National Laboratory, Los Alamos, NM 87544 (United States); Mazoyer, J. [LESIA, Observatoire de Paris, Meudon (France); Melikechi, N. [Delaware State University, Dover, DE (United States); Meslin, P.-Y. [Institut de Recherche en Astrophysique et Planetologie, Toulouse (France); Newsom, H. [University of New Mexico, Albuquerque, NM (United States); and others

    2013-04-01

    The ChemCam instrument package on the Mars Science Laboratory rover, Curiosity, is the first planetary science instrument to employ laser-induced breakdown spectroscopy (LIBS) to determine the compositions of geological samples on another planet. Pre-processing of the spectra involves subtracting the ambient light background, removing noise, removing the electron continuum, calibrating for the wavelength, correcting for the variable distance to the target, and applying a wavelength-dependent correction for the instrument response. Further processing of the data uses multivariate and univariate comparisons with a LIBS spectral library developed prior to launch as well as comparisons with several on-board standards post-landing. The level-2 data products include semi-quantitative abundances derived from partial least squares regression. A LIBS spectral library was developed using 69 rock standards in the form of pressed powder disks, glasses, and ceramics to minimize heterogeneity on the scale of the observation (350–550 μm dia.). The standards covered typical compositional ranges of igneous materials and also included sulfates, carbonates, and phyllosilicates. The provenance and elemental and mineralogical compositions of these standards are described. Spectral characteristics of this data set are presented, including the size distribution and integrated irradiances of the plasmas, and a proxy for plasma temperature as a function of distance from the instrument. Two laboratory-based clones of ChemCam reside in Los Alamos and Toulouse for the purpose of adding new spectra to the database as the need arises. Sensitivity to differences in wavelength correlation to spectral channels and spectral resolution has been investigated, indicating that spectral registration needs to be within half a pixel and resolution needs to match within 1.5 to 2.6 pixels. Absolute errors are tabulated for derived compositions of each major element in each standard using PLS regression

  12. CERN Radiation Protection (RP) calibration facilities

    CERN Document Server

    AUTHOR|(CDS)2082069; Macián-Juan, Rafael

    Radiation protection calibration facilities are essential to ensure the correct operation of radiation protection instrumentation. Calibrations are performed in specific radiation fields according to the type of instrument to be calibrated: neutrons, photons, X-rays, beta and alpha particles. Some of the instruments are also tested in mixed radiation fields as often encountered close to high-energy particle accelerators. Moreover, calibration facilities are of great importance to evaluate the performance of prototype detectors; testing and measuring the response of a prototype detector to well-known and -characterized radiation fields contributes to improving and optimizing its design and capabilities. The CERN Radiation Protection group is in charge of performing the regular calibrations of all CERN radiation protection devices; these include operational and passive dosimeters, neutron and photon survey-meters, and fixed radiation detectors to monitor the ambient dose equivalent, H*(10), inside CERN accelera...

  13. Radiation Calibration Measurements

    International Nuclear Information System (INIS)

    Omondi, C.

    2017-01-01

    KEBS Radiation Dosimetry mandate are: Custodian of Kenya Standards on Ionizing radiation, Ensure traceability to International System (SI ) and Calibration radiation equipment. RAF 8/040 on Radioisotope applications for troubleshooting and optimizing industrial process established Radiotracer Laboratory objective is to introduce and implement radiotracer technique for problem solving of industrial challenges. Gamma ray scanning technique applied is to Locate blockages, Locate liquid in vapor lines, Locate areas of lost refractory or lining in a pipe and Measure flowing densities. Equipment used for diagnostic and radiation protection must be calibrated to ensure Accuracy and Traceability

  14. Field calibration of cup anemometers

    Energy Technology Data Exchange (ETDEWEB)

    Kristensen, L.; Jensen, G.; Hansen, A.; Kirkegaard, P.

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statistical significance of the calibration expressions. It is concluded that the method has the advantage that many anemometers can be calibrated accurately with a minimum of work and cost. The obvious disadvantage is that the calibration of a set of anemometers may take more than one month in order to have wind speeds covering a sufficiently large magnitude range in a wind direction sector where we can be sure that the instruments are exposed to identical, simultaneous wind flows. Another main conclusion is that statistical uncertainty must be carefully evaluated since the individual 10 minute wind-speed averages are not statistically independent. (au)

  15. Automated system for calibration and control of the CHSPP-800 multichannel γ detector parameters

    International Nuclear Information System (INIS)

    Avvakumov, N.A.; Belikov, N.I.; Goncharenko, Yu.M.

    1987-01-01

    An automated system for adjustment, calibration and control of total absorption Cherenkov spectrometer is described. The system comprises a mechanical platform, capable of moving in two mutually perpendicular directions; movement detectors and limit switches; power unit, automation unit with remote control board. The automated system can operate both in manual control regime with coordinate control by a digital indicator, and in operation regime with computer according to special programs. The platform mounting accuracy is ± 0.1 mm. Application of the automated system has increased the rate of the course of the counter adjustment works 3-5 times

  16. Code Calibration as a Decision Problem

    DEFF Research Database (Denmark)

    Sørensen, John Dalsgaard; Kroon, I. B.; Faber, Michael Havbro

    1993-01-01

    Calibration of partial coefficients for a class of structures where no code exists is considered. The partial coefficients are determined such that the difference between the reliability for the different structures in the class considered and a target reliability level is minimized. Code...... calibration on a decision theoretical basis is discussed. Results from code calibration for rubble mound breakwater designs are shown....

  17. Calibration method of radiation monitoring system at TQNPC

    International Nuclear Information System (INIS)

    Liu Zhengshan; Zhang Qingli; Liu Jinjin; Miao Yuxing; Geng Lixin; Zhuang Yun; Dong Jianfeng; He Change

    2009-04-01

    The calibration methods and calibration device for standard monitor of radioactive particulate, iodine, noble gas and so on are not yet set up at home. On consideration of the present situation of the radiation monitoring system at the Third Qinshan Nuclear Power Co. Ltd., we have studied the calibration method of these radiation monitoring instruments used for measuring the waste liquid, particulate, iodine and noble gas produced during the operation of nuclear reactor. Through the check against these instruments during the No. 202 and No. 103 overhaul, we got initially the method of the calibration and obtained the transfer coefficient of calibration when secondary solid sources are used for calibration. Through the testing and calibration, the credibility of the radiation monitoring system is enhanced. And at the same time, the problems existing in the calibration are discussed. (authors)

  18. Dosemeter calibration in X-ray and in cobalt-60

    International Nuclear Information System (INIS)

    Silva, T.A. da

    1988-01-01

    Some tests about quality security for clinical dosemeter calibration in secondary standard dosimetry laboratory are described. The tests in gama calibration system, in X-ray calibration, in secondary standard dosimeter, in the dosemeter that will be calibrated, during the calibration and after the calibration are shown. (C.G.C.) [pt

  19. Exploring a Black Body Source as an Absolute Radiometric Calibration Standard and Comparison with a NIST Traced Lamp Standard

    Science.gov (United States)

    Green, Robert O.; Chrien, Thomas; Sarture, Chuck

    2001-01-01

    Radiometric calibration of the Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) is required for the scientific research and application objectives pursued with the spectroscopic measurements. Specifically calibration is required for: inter-comparison of AVIRIS data measured at different locations and at different times; analysis of AVIRIS data with data measured by other instruments; and analysis of AVIRIS data in conjunction with computer models. The primary effect of radiometric calibration is conversion of AVIRIS instrument response values (digitized numbers, or DN) to units of absolute radiance. For example, a figure shows the instrument response spectrum measured by AVIRIS over a portion of Rogers Dry Lake, California, and another figure shows the same spectrum calibrated to radiance. Only the calibrated spectrum may be quantitatively analyzed for science research and application objectives. Since the initial development of the AVIRIS instrument-radiometric calibration has been based upon a 1000-W irradiance lamp with a calibration traced to the National Institute of Standards and Technology (NIST). There are several advantages to this irradiance-lamp calibration approach. First, the considerable effort of NIST backs up the calibration. Second, by changing the distance to the lamp, the output can closely span the radiance levels measured by AVIRIS. Third, this type of standard is widely used. Fourth, these calibrated lamps are comparatively inexpensive. Conversely, there are several disadvantages to this approach as well. First, the lamp is not a primary standard. Second, the lamp output characteristics may change in an unknown manner through time. Third, it is difficult to assess, constrain, or improve the calibration uncertainty delivered with the lamp. In an attempt to explore the effect and potentially address some of these disadvantages a set of analyses and measurements comparing an irradiance lamp with a black-body source have been completed

  20. Calibration of thermoluminiscent materials

    International Nuclear Information System (INIS)

    Bos, A.J.J.

    1989-07-01

    In this report the relation between exposure and absorbed radiation dose in various materials is represented, on the base of recent data. With the help of this a calibration procedure for thermoluminescent materials, adapted to the IRI radiation standard is still the exposure in rontgen. In switching to the air kerma standard the calibration procedure will have to be adapted. (author). 6 refs.; 4 tabs

  1. Electromagnetic Calorimeter Calibration with $\\pi^{0}$

    CERN Multimedia

    Puig Navarro, A

    2009-01-01

    Several methods can be used in order to achieve precise calibration of the LHCb Electromagnetic Calorimeter (ECAL) once reasonable cell equalization has been reached. At low transverse energy, the standard calibration procedure is an iterative method based on the fit of the $\\gamma\\gamma$ invariant mass distribution for each cell of the decay $\\pi^{0}\\to\\gamma\\gamma$ with resolved photons. A new technique for generating the combinatorial background of such decays directly from data has been developed. Knowledge of the background could allow an alternative calibration method based on a event by event fit of the same $\\gamma\\gamma$ invariant mass distribution where contributions from groups of cells are considered in a single fit. The background generation procedure and this possible new calibration method are presented in this poster, in addition to an overview of the LHCb Calorimetry system and ECAL calibration techniques.

  2. The ATLAS Electromagnetic Calorimeter Calibration Workshop

    CERN Multimedia

    Hong Ma; Isabelle Wingerter

    The ATLAS Electromagnetic Calorimeter Calibration Workshop took place at LAPP-Annecy from the 1st to the 3rd of October; 45 people attended the workshop. A detailed program was setup before the workshop. The agenda was organised around very focused presentations where questions were raised to allow arguments to be exchanged and answers to be proposed. The main topics were: Electronics calibration Handling of problematic channels Cluster level corrections for electrons and photons Absolute energy scale Streams for calibration samples Calibration constants processing Learning from commissioning Forty-five people attended the workshop. The workshop was on the whole lively and fruitful. Based on years of experience with test beam analysis and Monte Carlo simulation, and the recent operation of the detector in the commissioning, the methods to calibrate the electromagnetic calorimeter are well known. Some of the procedures are being exercised in the commisssioning, which have demonstrated the c...

  3. Two calibration procedures for a gyroscope-free inertial measurement system based on a double-pendulum apparatus

    Science.gov (United States)

    Cappa, P.; Patanè, F.; Rossi, S.

    2008-05-01

    This paper presents a novel calibration algorithm to be used with a gyro-free inertial measurement unit (GF-IMU) based on the use of linear accelerometers (AC). The analytical approach can be implemented in two calibration procedures. The first procedure (P-I) is articulated in the conduction of a static trial, to compute the sensitivity and the direction of the sensing axis of each AC, followed by a dynamic trial, to determine the AC locations. By contrast, the latter procedure (P-II) consists in the calculation of the previously indicated calibration parameters by means of a dynamic trial only. The feasibility of the two calibration procedures has been investigated by testing two GF-IMUs, equipped with ten and six bi-axial linear ACs, with an ad hoc instrumented double-pendulum apparatus. P-I and P-II were compared to a calibration procedure used as a reference (P-REF), which incorporates the AC positions measured with an optoelectronic system. The experimental results we present in this paper demonstrate that (i) P-I is able to determine the calibration parameters of the AC array with a higher accuracy than P-II; (ii) consequently, the errors associated with translational (a0 - g) and rotational (\\dot{\\bm \\omega }) acceleration components for the two GF-IMUs are significantly greater using P-II than P-I and (iii) the errors in (a0 - g) and \\dot{\\bm \\omega } obtained with P-I are comparable with the ones obtainable by using P-REF. Thus, the proposed novel algorithm used in P-I, in conjunction with the double-pendulum apparatus, can be globally considered a viable tool in GF-IMU calibration.

  4. Calibrated Properties Model

    International Nuclear Information System (INIS)

    Ghezzehej, T.

    2004-01-01

    The purpose of this model report is to document the calibrated properties model that provides calibrated property sets for unsaturated zone (UZ) flow and transport process models (UZ models). The calibration of the property sets is performed through inverse modeling. This work followed, and was planned in, ''Technical Work Plan (TWP) for: Unsaturated Zone Flow Analysis and Model Report Integration'' (BSC 2004 [DIRS 169654], Sections 1.2.6 and 2.1.1.6). Direct inputs to this model report were derived from the following upstream analysis and model reports: ''Analysis of Hydrologic Properties Data'' (BSC 2004 [DIRS 170038]); ''Development of Numerical Grids for UZ Flow and Transport Modeling'' (BSC 2004 [DIRS 169855]); ''Simulation of Net Infiltration for Present-Day and Potential Future Climates'' (BSC 2004 [DIRS 170007]); ''Geologic Framework Model'' (GFM2000) (BSC 2004 [DIRS 170029]). Additionally, this model report incorporates errata of the previous version and closure of the Key Technical Issue agreement TSPAI 3.26 (Section 6.2.2 and Appendix B), and it is revised for improved transparency

  5. On Inertial Body Tracking in the Presence of Model Calibration Errors.

    Science.gov (United States)

    Miezal, Markus; Taetz, Bertram; Bleser, Gabriele

    2016-07-22

    In inertial body tracking, the human body is commonly represented as a biomechanical model consisting of rigid segments with known lengths and connecting joints. The model state is then estimated via sensor fusion methods based on data from attached inertial measurement units (IMUs). This requires the relative poses of the IMUs w.r.t. the segments-the IMU-to-segment calibrations, subsequently called I2S calibrations-to be known. Since calibration methods based on static poses, movements and manual measurements are still the most widely used, potentially large human-induced calibration errors have to be expected. This work compares three newly developed/adapted extended Kalman filter (EKF) and optimization-based sensor fusion methods with an existing EKF-based method w.r.t. their segment orientation estimation accuracy in the presence of model calibration errors with and without using magnetometer information. While the existing EKF-based method uses a segment-centered kinematic chain biomechanical model and a constant angular acceleration motion model, the newly developed/adapted methods are all based on a free segments model, where each segment is represented with six degrees of freedom in the global frame. Moreover, these methods differ in the assumed motion model (constant angular acceleration, constant angular velocity, inertial data as control input), the state representation (segment-centered, IMU-centered) and the estimation method (EKF, sliding window optimization). In addition to the free segments representation, the optimization-based method also represents each IMU with six degrees of freedom in the global frame. In the evaluation on simulated and real data from a three segment model (an arm), the optimization-based method showed the smallest mean errors, standard deviations and maximum errors throughout all tests. It also showed the lowest dependency on magnetometer information and motion agility. Moreover, it was insensitive w.r.t. I2S position and

  6. 14 MeV calibration of JET neutron detectors—phase 1: calibration and characterization of the neutron source

    Science.gov (United States)

    Batistoni, P.; Popovichev, S.; Cufar, A.; Ghani, Z.; Giacomelli, L.; Jednorog, S.; Klix, A.; Lilley, S.; Laszynska, E.; Loreti, S.; Packer, L.; Peacock, A.; Pillon, M.; Price, R.; Rebai, M.; Rigamonti, D.; Roberts, N.; Tardocchi, M.; Thomas, D.; Contributors, JET

    2018-02-01

    In view of the planned DT operations at JET, a calibration of the JET neutron monitors at 14 MeV neutron energy is needed using a 14 MeV neutron generator deployed inside the vacuum vessel by the JET remote handling system. The target accuracy of this calibration is  ±10% as also required by ITER, where a precise neutron yield measurement is important, e.g. for tritium accountancy. To achieve this accuracy, the 14 MeV neutron generator selected as the calibration source has been fully characterised and calibrated prior to the in-vessel calibration of the JET monitors. This paper describes the measurements performed using different types of neutron detectors, spectrometers, calibrated long counters and activation foils which allowed us to obtain the neutron emission rate and the anisotropy of the neutron generator, i.e. the neutron flux and energy spectrum dependence on emission angle, and to derive the absolute emission rate in 4π sr. The use of high resolution diamond spectrometers made it possible to resolve the complex features of the neutron energy spectra resulting from the mixed D/T beam ions reacting with the D/T nuclei present in the neutron generator target. As the neutron generator is not a stable neutron source, several monitoring detectors were attached to it by means of an ad hoc mechanical structure to continuously monitor the neutron emission rate during the in-vessel calibration. These monitoring detectors, two diamond diodes and activation foils, have been calibrated in terms of neutrons/counts within  ±5% total uncertainty. A neutron source routine has been developed, able to produce the neutron spectra resulting from all possible reactions occurring with the D/T ions in the beam impinging on the Ti D/T target. The neutron energy spectra calculated by combining the source routine with a MCNP model of the neutron generator have been validated by the measurements. These numerical tools will be key in analysing the results from the in

  7. Energy Calibration of a Silicon-Strip Detector for Photon-Counting Spectral CT by Direct Usage of the X-ray Tube Spectrum

    Science.gov (United States)

    Liu, Xuejin; Chen, Han; Bornefalk, Hans; Danielsson, Mats; Karlsson, Staffan; Persson, Mats; Xu, Cheng; Huber, Ben

    2015-02-01

    The variation among energy thresholds in a multibin detector for photon-counting spectral CT can lead to ring artefacts in the reconstructed images. Calibration of the energy thresholds can be used to achieve homogeneous threshold settings or to develop compensation methods to reduce the artefacts. We have developed an energy-calibration method for the different comparator thresholds employed in a photon-counting silicon-strip detector. In our case, this corresponds to specifying the linear relation between the threshold positions in units of mV and the actual deposited photon energies in units of keV. This relation is determined by gain and offset values that differ for different detector channels due to variations in the manufacturing process. Typically, the calibration is accomplished by correlating the peak positions of obtained pulse-height spectra to known photon energies, e.g. with the aid of mono-energetic x rays from synchrotron radiation, radioactive isotopes or fluorescence materials. Instead of mono-energetic x rays, the calibration method presented in this paper makes use of a broad x-ray spectrum provided by commercial x-ray tubes. Gain and offset as the calibration parameters are obtained by a regression analysis that adjusts a simulated spectrum of deposited energies to a measured pulse-height spectrum. Besides the basic photon interactions such as Rayleigh scattering, Compton scattering and photo-electric absorption, the simulation takes into account the effect of pulse pileup, charge sharing and the electronic noise of the detector channels. We verify the method for different detector channels with the aid of a table-top setup, where we find the uncertainty of the keV-value of a calibrated threshold to be between 0.1 and 0.2 keV.

  8. Calibration of hydrometers

    Science.gov (United States)

    Lorefice, Salvatore; Malengo, Andrea

    2006-10-01

    After a brief description of the different methods employed in periodic calibration of hydrometers used in most cases to measure the density of liquids in the range between 500 kg m-3 and 2000 kg m-3, particular emphasis is given to the multipoint procedure based on hydrostatic weighing, known as well as Cuckow's method. The features of the calibration apparatus and the procedure used at the INRiM (formerly IMGC-CNR) density laboratory have been considered to assess all relevant contributions involved in the calibration of different kinds of hydrometers. The uncertainty is strongly dependent on the kind of hydrometer; in particular, the results highlight the importance of the density of the reference buoyant liquid, the temperature of calibration and the skill of operator in the reading of the scale in the whole assessment of the uncertainty. It is also interesting to realize that for high-resolution hydrometers (division of 0.1 kg m-3), the uncertainty contribution of the density of the reference liquid is the main source of the total uncertainty, but its importance falls under about 50% for hydrometers with a division of 0.5 kg m-3 and becomes somewhat negligible for hydrometers with a division of 1 kg m-3, for which the reading uncertainty is the predominant part of the total uncertainty. At present the best INRiM result is obtained with commercially available hydrometers having a scale division of 0.1 kg m-3, for which the relative uncertainty is about 12 × 10-6.

  9. Color calibration and color-managed medical displays: does the calibration method matter?

    Science.gov (United States)

    Roehrig, Hans; Rehm, Kelly; Silverstein, Louis D.; Dallas, William J.; Fan, Jiahua; Krupinski, Elizabeth A.

    2010-02-01

    Our laboratory has investigated the efficacy of a suite of color calibration and monitor profiling packages which employ a variety of color measurement sensors. Each of the methods computes gamma correction tables for the red, green and blue color channels of a monitor that attempt to: a) match a desired luminance range and tone reproduction curve; and b) maintain a target neutral point across the range of grey values. All of the methods examined here produce International Color Consortium (ICC) profiles that describe the color rendering capabilities of the monitor after calibration. Color profiles incorporate a transfer matrix that establishes the relationship between RGB driving levels and the International Commission on Illumination (CIE) XYZ (tristimulus) values of the resulting on-screen color; the matrix is developed by displaying color patches of known RGB values on the monitor and measuring the tristimulus values with a sensor. The number and chromatic distribution of color patches varies across methods and is usually not under user control. In this work we examine the effect of employing differing calibration and profiling methods on rendition of color images. A series of color patches encoded in sRGB color space were presented on the monitor using color-management software that utilized the ICC profile produced by each method. The patches were displayed on the calibrated monitor and measured with a Minolta CS200 colorimeter. Differences in intended and achieved luminance and chromaticity were computed using the CIE DE2000 color-difference metric, in which a value of ▵E = 1 is generally considered to be approximately one just noticeable difference (JND) in color. We observed between one and 17 JND's for individual colors, depending on calibration method and target.

  10. Laboratory implantation for well type ionization chambers calibration

    International Nuclear Information System (INIS)

    Vianello, E.A.; Dias, D.J.; Almeida, C.E. de

    1998-01-01

    The Radiological Science Laboratory is implanting a service for calibration of well type chambers by IAEA training program. The kerma rate in the air (mu Gy/h) of the linear Cs-137 reference source CDCS-J4 have been determined using a well type chamber Standard Imaging HDR-1000 model, which have been calibrated at Secondary Standard Laboratory Calibration of IAEA, whereas two HDR-1000 Plus chambers were calibrated too, following the same standards. The results were compared with Wisconsin University calibration certification and has demonstrated that well type ionization chamber calibration can be used in brachytherapy for several kinds of radionuclides. (Author)

  11. Reactor operations, inspection and maintenance. PNGS Calibration Program

    International Nuclear Information System (INIS)

    Lopez, E.

    1997-01-01

    The PNGS Calibration Program is being implemented as a response to various concerns identified in recent PEER evaluations and AECB audits. Identified areas of concern were the approach to instrument calibration of Special Safety Systems (SSS). The implementation of a calibration program is a significant improvement in operating practices. A systematic and comprehensive approach to calibration of instrumentation will improve the quality of operation of the plant with a positive contribution to PNGS safety of operation and economic objectives. This paper describes the strategy to implement the proposed calibration program and describes its calibration data requirements. (DM)

  12. Cross-correlation redshift calibration without spectroscopic calibration samples in DES Science Verification Data

    Science.gov (United States)

    Davis, C.; Rozo, E.; Roodman, A.; Alarcon, A.; Cawthon, R.; Gatti, M.; Lin, H.; Miquel, R.; Rykoff, E. S.; Troxel, M. A.; Vielzeuf, P.; Abbott, T. M. C.; Abdalla, F. B.; Allam, S.; Annis, J.; Bechtol, K.; Benoit-Lévy, A.; Bertin, E.; Brooks, D.; Buckley-Geer, E.; Burke, D. L.; Carnero Rosell, A.; Carrasco Kind, M.; Carretero, J.; Castander, F. J.; Crocce, M.; Cunha, C. E.; D'Andrea, C. B.; da Costa, L. N.; Desai, S.; Diehl, H. T.; Doel, P.; Drlica-Wagner, A.; Fausti Neto, A.; Flaugher, B.; Fosalba, P.; Frieman, J.; García-Bellido, J.; Gaztanaga, E.; Gerdes, D. W.; Giannantonio, T.; Gruen, D.; Gruendl, R. A.; Gutierrez, G.; Honscheid, K.; Jain, B.; James, D. J.; Jeltema, T.; Krause, E.; Kuehn, K.; Kuhlmann, S.; Kuropatkin, N.; Lahav, O.; Li, T. S.; Lima, M.; March, M.; Marshall, J. L.; Martini, P.; Melchior, P.; Ogando, R. L. C.; Plazas, A. A.; Romer, A. K.; Sanchez, E.; Scarpine, V.; Schindler, R.; Schubnell, M.; Sevilla-Noarbe, I.; Smith, M.; Soares-Santos, M.; Sobreira, F.; Suchyta, E.; Swanson, M. E. C.; Tarle, G.; Thomas, D.; Vikram, V.; Walker, A. R.; Wechsler, R. H.

    2018-06-01

    Galaxy cross-correlations with high-fidelity redshift samples hold the potential to precisely calibrate systematic photometric redshift uncertainties arising from the unavailability of complete and representative training and validation samples of galaxies. However, application of this technique in the Dark Energy Survey (DES) is hampered by the relatively low number density, small area, and modest redshift overlap between photometric and spectroscopic samples. We propose instead using photometric catalogues with reliable photometric redshifts for photo-z calibration via cross-correlations. We verify the viability of our proposal using redMaPPer clusters from the Sloan Digital Sky Survey (SDSS) to successfully recover the redshift distribution of SDSS spectroscopic galaxies. We demonstrate how to combine photo-z with cross-correlation data to calibrate photometric redshift biases while marginalizing over possible clustering bias evolution in either the calibration or unknown photometric samples. We apply our method to DES Science Verification (DES SV) data in order to constrain the photometric redshift distribution of a galaxy sample selected for weak lensing studies, constraining the mean of the tomographic redshift distributions to a statistical uncertainty of Δz ˜ ±0.01. We forecast that our proposal can, in principle, control photometric redshift uncertainties in DES weak lensing experiments at a level near the intrinsic statistical noise of the experiment over the range of redshifts where redMaPPer clusters are available. Our results provide strong motivation to launch a programme to fully characterize the systematic errors from bias evolution and photo-z shapes in our calibration procedure.

  13. Status report of the ATR SPING-3A and detailed calibration guideline

    International Nuclear Information System (INIS)

    Koeppen, L.D.; Rogers, J.W.

    1985-12-01

    The Radiation Measurements Laboratory (RML) onducted testing and an evaluation of the ATR SPING-3 after installation of the new ''A'' series software (firmware) and retrofit kit. The purpose of this report is to provide the ATR SPING-3A users with a document describing the present status of the unit, how the system was tested, a suggested calibration guideline, how the system should be operated for the ATR application, and why it should be used in this manner

  14. Field calibration of cup anemometers

    DEFF Research Database (Denmark)

    Kristensen, L.; Jensen, G.; Hansen, A.

    2001-01-01

    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statisticalsignificance of the calibration expressions...

  15. Calibration procedure of Hukseflux SR25 to Establish the Diffuse Reference for the Outdoor Broadband Radiometer Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Reda, Ibrahim M. [National Renewable Energy Laboratory (NREL), Golden, CO (United States); Andreas, Afshin M. [National Renewable Energy Laboratory (NREL), Golden, CO (United States)

    2017-08-01

    Accurate pyranometer calibrations, traceable to internationally recognized standards, are critical for solar irradiance measurements. One calibration method is the component summation method, where the pyranometers are calibrated outdoors under clear sky conditions, and the reference global solar irradiance is calculated as the sum of two reference components, the diffuse horizontal and subtended beam solar irradiances. The beam component is measured with pyrheliometers traceable to the World Radiometric Reference, while there is no internationally recognized reference for the diffuse component. In the absence of such a reference, we present a method to consistently calibrate pyranometers for measuring the diffuse component. The method is based on using a modified shade/unshade method and a pyranometer with less than 0.5 W/m2 thermal offset. The calibration result shows that the responsivity of Hukseflux SR25 pyranometer equals 10.98 uV/(W/m2) with +/-0.86 percent uncertainty.

  16. Standard-target calibration of an acoustic backscatter system

    Science.gov (United States)

    Foote, Kenneth G.; Martini, Marinna A.

    2010-01-01

    The standard-target method used to calibrate scientific echo sounders and other scientific sonars by a single, solid elastic sphere is being adapted to acoustic backscatter (ABS) systems. Its first application, to the AQUAscat 1000, is described. The on-axis sensitivity and directional properties of transducer beams at three operating frequencies, nominally 1, 2.5, and 4 MHz, have been determined using a 10-mm-diameter sphere of tungsten carbide with 6% cobalt binder. Preliminary results are reported for the 1-MHz transducer. Their application to measurements of suspended sediment made in situ with the same device is described. This will enable the data to be expressed directly in physical units of volume backscattering.

  17. The third flight of the Colorado high-resolution echelle stellar spectrograph (CHESS): improvements, calibrations, and preliminary results

    Science.gov (United States)

    Kruczek, Nicholas; Nell, Nicholas; France, Kevin; Hoadley, Keri; Fleming, Brian; Kane, Robert; Ulrich, Stefan; Egan, Arika; Beatty, Dawson

    2017-08-01

    In this proceeding, we describe the scientific motivation and technical development of the Colorado HighResolution Echelle Stellar Spectrograph (CHESS), focusing on the hardware advancements and testing of components for the third launch of the payload (CHESS-3). CHESS is a far ultraviolet rocket-borne instrument designed to study the atomic-to-molecular transitions within translucent cloud regions in the interstellar medium. CHESS is an objective echelle spectrograph, which uses a mechanically-ruled echelle and a powered (f/12.4) crossdispersing grating, and is designed to achieve a resolving power R > 100,000 over the bandpass λλ 1000-1600 Å. Results from final efficiency and reflectivity measurements for the optical components of CHESS-3 are presented. An important role of sounding rocket experiments is the testing and verification of the space flight capabilities of experimental technologies. CHESS-3 utilizes a 40mm-diameter cross-strip anode microchannel plate detector fabricated by Sensor Sciences LLC, capable of achieving high spatial resolution and a high global count rate (˜1 MHz). We present pre-flight laboratory spectra and calibration results, including wavelength solution and resolving power of the instrument. The fourth launch of CHESS (CHESS-4) will demonstrate a δ-doped CCD, assembled in collaboration with the Microdevices Laboratory at JPL and Arizona State University. In support of CHESS-4, the CHESS-3 payload included a photomultiplier tube, used as a secondary confirmation of the optical alignment of the payload during flight. CHESS-3 launched on 26 June 2017 aboard NASA/CU sounding rocket mission 36.323 UG. We present initial flight results for the CHESS-3 observation of the β1 Scorpii sightline.

  18. Cryogenic Thermometer Calibration Facility at CERN

    CERN Document Server

    Balle, C; Thermeau, J P

    1998-01-01

    A cryogenic thermometer calibration facility has been designed and is being commissioned in preparation for the very stringent requirements on the temperature control of the LHC superconducting magnets. The temperature is traceable in the 1.5 to 30 K range to standards maintained in a national metrological laboratory by using a set of Rhodium-Iron temperature sensors of metrological quality. The calibration facility is designed for calibrating simultaneously 60 industrial cryogenic thermometers in the 1.5 K to 300 K temperature range, a thermometer being a device that includes both a temperature sensor and the wires heat-intercept. The thermometers can be calibrated in good and degraded vacuum or immersed in the surrounding fluid and at different Joule self-heating conditions that match those imposed by signal conditioners used in large cryogenic machinery. The calibration facility can be operated in an automatic mode and all the control and safety routines are handled by a Programmable Logic Controller (PLC)...

  19. On-line monitoring for calibration reduction

    International Nuclear Information System (INIS)

    Hoffmann, M.

    2005-09-01

    On-Line Monitoring evaluates instrument channel performance by assessing its consistency with other plant indications. Elimination or reduction of unnecessary field calibrations can reduce associated labour costs, reduce personnel radiation exposure, and reduce the potential for calibration errors. On-line calibration monitoring is an important technique to implement a state-based maintenance approach and reduce unnecessary field calibrations. In this report we will look at how the concept is currently applied in the industry and what the arising needs are as it becomes more commonplace. We will also look at the PEANO System, a tool developed by the Halden Project to perform signal validation and on-line calibration monitoring. Some issues will be identified that are being addressed in the further development of these tools to better serve the future needs of the industry in this area. An outline for how to improve these points and which aspects should be taken into account is described in detail. (Author)

  20. Hydrology and sediment yield calibration for the Barasona reservoir catchment (Spain) using SWAT

    Science.gov (United States)

    Palazón, Leticia; Navas, Ana

    2013-04-01

    Hydrological and soil erosion models, as Soil and Water Assessment Tool (SWAT), have become very useful tools and increasingly serve as vital components of integrated environmental assessments that provide information outside of direct field experiments and causal observation. The purpose of this study was to improve the calibration of SWAT model to use it in an alpine catchment as a simulator of processes related to water quality and soil erosion. SWAT is spatially semi-distributed, agro-hydrological model that operates on a daily time step (as a minimum) at basin scale. It is designed to predict the impact of management on water, sediment and agricultural chemical yields in ungaged catchments. SWAT provides physically based algorithms as an option to define many of the important components of the hydrologic cycle. The input requirements of the model are used to describe the climate, soil properties, topography, vegetation, and land management practices. SWAT analyzes small or large catchments by discretising into sub-basins, which are then further subdivided into hydrological response units (HRUs) with homogeneous land use, soil type and slope. SWAT model (SWAT2009) coupled with a GIS interface (ArcSWAT), was applied to the Barasona reservoir catchment located in the central Spanish Pyrenees. The 1509 km2 agro-forestry catchment presents a mountain type climate, an altitudinal range close to 3000 meters and a precipitation variation close to 1000 mm/km. The mountainous characteristics of the catchment, in addition to the scarcity of climate data in the region, require specific calibration for some processes. Snowfall and snowmelt are significant processes in the hydrologic regime of the area and were calibrated in a previous work. In this work some of the challenges of the catchment to model with SWAT which affected the hydrology and the sediment yield simulation were performed as improvement of the previous calibration. Two reservoirs, a karst system which

  1. Linear Calibration – Is It so Simple?

    International Nuclear Information System (INIS)

    Arsova, Diana; Babanova, Sofia; Mandjukov, Petko

    2009-01-01

    Calibration procedure is an important part of instrumental analysis. Usually it is not the major uncertainty source in whole analytical procedure. However, improper calibration might cause a significant bias of the analytical results from the real (certified) value. Standard Gaussian linear regression is the most frequently used mathematical approach for estimation of calibration function parameters. In the present article are discussed some not quite popular, but highly recommended in certain cases methods for parameter estimation, such as: weighted regression, orthogonal regression, robust regression, bracketing calibration etc. Some useful approximations are also presented. Special attention is paid to the statistical criteria which to be used for selection of proper calibration model. Standard UV-VIS spectrometric procedure for determination of phosphates in water was used as a practical example. Several different approaches for estimation of the contribution of calibration to the general un-certainty of the analytical result are presented and compared

  2. Calibration of Photon Sources for Brachytherapy

    Science.gov (United States)

    Rijnders, Alex

    Source calibration has to be considered an essential part of the quality assurance program in a brachytherapy department. Not only it will ensure that the source strength value used for dose calculation agrees within some predetermined limits to the value stated on the source certificate, but also it will ensure traceability to international standards. At present calibration is most often still given in terms of reference air kerma rate, although calibration in terms of absorbed dose to water would be closer to the users interest. It can be expected that in a near future several standard laboratories will be able to offer this latter service, and dosimetry protocols will have to be adapted in this way. In-air measurement using ionization chambers (e.g. a Baldwin—Farmer ionization chamber for 192Ir high dose rate HDR or pulsed dose rate PDR sources) is still considered the method of choice for high energy source calibration, but because of their ease of use and reliability well type chambers are becoming more popular and are nowadays often recommended as the standard equipment. For low energy sources well type chambers are in practice the only equipment available for calibration. Care should be taken that the chamber is calibrated at the standard laboratory for the same source type and model as used in the clinic, and using the same measurement conditions and setup. Several standard laboratories have difficulties to provide these calibration facilities, especially for the low energy seed sources (125I and 103Pd). Should a user not be able to obtain properly calibrated equipment to verify the brachytherapy sources used in his department, then at least for sources that are replaced on a regular basis, a consistency check program should be set up to ensure a minimal level of quality control before these sources are used for patient treatment.

  3. Guidelines on calibration of neutron measuring devices

    International Nuclear Information System (INIS)

    Burger, G.

    1988-01-01

    The International Atomic Energy Agency and the World Health Organization have agreed to establish an IAEA/WHO Network of Secondary Standard Dosimetry Laboratories (SSDLs) in order to improve accuracy in applied radiation dosimetry throughout the world. These SSDLs must be equipped with, and maintain, secondary standard instruments, which have been calibrated against primary standards, and must be nominated by their governments for membership of the network. The majority of the existing SSDLs were established primarily to work with photon radiation (X-rays and gamma rays). Neutron sources are, however, increasingly being applied in industrial processes, research, nuclear power development and radiation biology and medicine. Thus, it is desirable that the SSDLs in countries using neutron sources on a regular basis should also fulfil the minimum requirements to calibrate neutron measuring devices. It is the primary purpose of this handbook to provide guidance on calibration of instruments for radiation protection. A calibration laboratory should also be in a position to calibrate instrumentation being used for the measurement of kerma and absorbed dose and their corresponding rates. This calibration is generally done with photons. In addition, since each neutron field is usually contaminated by photons produced in the source or by scatter in the surrounding media, neutron protection instrumentation has to be tested with respect to its intrinsic photon response. The laboratory will therefore need to possess equipment for photon calibration. This publication deals primarily with methods of applying radioactive neutron sources for calibration of instrumentation, and gives an indication of the space, manpower and facilities needed to fulfil the minimum requirements of a calibration laboratory for neutron work. It is intended to serve as a guide for centres about to start on neutron dosimetry standardization and calibration. 94 refs, 8 figs, 12 tabs

  4. Out of lab calibration of a rotating 2D scanner for 3D mapping

    Science.gov (United States)

    Koch, Rainer; Böttcher, Lena; Jahrsdörfer, Maximilian; Maier, Johannes; Trommer, Malte; May, Stefan; Nüchter, Andreas

    2017-06-01

    Mapping is an essential task in mobile robotics. To fulfil advanced navigation and manipulation tasks a 3D representation of the environment is required. Applying stereo cameras or Time-of-flight cameras (TOF cameras) are one way to archive this requirement. Unfortunately, they suffer from drawbacks which makes it difficult to map properly. Therefore, costly 3D laser scanners are applied. An inexpensive way to build a 3D representation is to use a 2D laser scanner and rotate the scan plane around an additional axis. A 3D point cloud acquired with such a custom device consists of multiple 2D line scans. Therefore the scanner pose of each line scan need to be determined as well as parameters resulting from a calibration to generate a 3D point cloud. Using external sensor systems are a common method to determine these calibration parameters. This is costly and difficult when the robot needs to be calibrated outside the lab. Thus, this work presents a calibration method applied on a rotating 2D laser scanner. It uses a hardware setup to identify the required parameters for calibration. This hardware setup is light, small, and easy to transport. Hence, an out of lab calibration is possible. Additional a theoretical model was created to test the algorithm and analyse impact of the scanner accuracy. The hardware components of the 3D scanner system are an HOKUYO UTM-30LX-EW 2D laser scanner, a Dynamixel servo-motor, and a control unit. The calibration system consists of an hemisphere. In the inner of the hemisphere a circular plate is mounted. The algorithm needs to be provided with a dataset of a single rotation from the laser scanner. To achieve a proper calibration result the scanner needs to be located in the middle of the hemisphere. By means of geometric formulas the algorithms determine the individual deviations of the placed laser scanner. In order to minimize errors, the algorithm solves the formulas in an iterative process. First, the calibration algorithm was

  5. Bayesian calibration : past achievements and future challenges

    International Nuclear Information System (INIS)

    Christen, J.A.

    2001-01-01

    Due to variations of the radiocarbon content in the biosphere over time, radiocarbon determinations need to be calibrated to obtain calendar years. Over the past decade a series of researchers have investigated the possibility of using Bayesian statistics to calibrate radiocarbon determinations, the main feature being the inclusion of contextual information into the calibration process. This allows for a coherent calibration of groups of determinations arising from related contexts (stratigraphical layers, peat cores, cultural events, ect.). Moreover, the 'related contexts' are also dated, and not only the material radiocarbon dated itself. We review Bayesian Calibration and state some of its current challenges like: software development, prior specification, robustness, etc. (author). 14 refs., 4 figs

  6. Recommended inorganic chemicals for calibration

    International Nuclear Information System (INIS)

    Moody, J.R.; Greenberg, R.R.; Pratt, K.W.; Rains, T.C.

    1988-01-01

    All analytical techniques depend on the use of calibration chemicals to relate analyte concentration to an instrumental parameter. A fundamental component in the preparation of calibration solutions is the weighing of a pure chemical or metal before preparing a solution standard. The analyst must be assured that the purity, stoichiometry, and assay of the chemical are known. These terms have different meanings, and each has an important influence. This report is intended to assist the analyst in the selection and use of chemical standards for instrumental calibration. Purity, stoichiometry, and preparation of solutions for different purposes are discussed, and a critical evaluation of the best materials available for each element is presented for use in preparing solutions or calibration standards. Information on the chemical form, source, purity, drying, and appropriate precautions is given. In some cases, multiple sources or chemical forms are available. Certain radioactive elements, the transuranic elements, and the noble gases are not considered

  7. 21 CFR 862.3200 - Clinical toxicology calibrator.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Clinical toxicology calibrator. 862.3200 Section... (CONTINUED) MEDICAL DEVICES CLINICAL CHEMISTRY AND CLINICAL TOXICOLOGY DEVICES Clinical Toxicology Test Systems § 862.3200 Clinical toxicology calibrator. (a) Identification. A clinical toxicology calibrator is...

  8. Automatic multi-camera calibration for deployable positioning systems

    Science.gov (United States)

    Axelsson, Maria; Karlsson, Mikael; Rudner, Staffan

    2012-06-01

    Surveillance with automated positioning and tracking of subjects and vehicles in 3D is desired in many defence and security applications. Camera systems with stereo or multiple cameras are often used for 3D positioning. In such systems, accurate camera calibration is needed to obtain a reliable 3D position estimate. There is also a need for automated camera calibration to facilitate fast deployment of semi-mobile multi-camera 3D positioning systems. In this paper we investigate a method for automatic calibration of the extrinsic camera parameters (relative camera pose and orientation) of a multi-camera positioning system. It is based on estimation of the essential matrix between each camera pair using the 5-point method for intrinsically calibrated cameras. The method is compared to a manual calibration method using real HD video data from a field trial with a multicamera positioning system. The method is also evaluated on simulated data from a stereo camera model. The results show that the reprojection error of the automated camera calibration method is close to or smaller than the error for the manual calibration method and that the automated calibration method can replace the manual calibration.

  9. Sensor Calibration Design Based on D-Optimality Criterion

    Directory of Open Access Journals (Sweden)

    Hajiyev Chingiz

    2016-09-01

    Full Text Available In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM. As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.

  10. Data multiplexing in radio interferometric calibration

    Science.gov (United States)

    Yatawatta, Sarod; Diblen, Faruk; Spreeuw, Hanno; Koopmans, L. V. E.

    2018-03-01

    New and upcoming radio interferometers will produce unprecedented amount of data that demand extremely powerful computers for processing. This is a limiting factor due to the large computational power and energy costs involved. Such limitations restrict several key data processing steps in radio interferometry. One such step is calibration where systematic errors in the data are determined and corrected. Accurate calibration is an essential component in reaching many scientific goals in radio astronomy and the use of consensus optimization that exploits the continuity of systematic errors across frequency significantly improves calibration accuracy. In order to reach full consensus, data at all frequencies need to be calibrated simultaneously. In the SKA regime, this can become intractable if the available compute agents do not have the resources to process data from all frequency channels simultaneously. In this paper, we propose a multiplexing scheme that is based on the alternating direction method of multipliers with cyclic updates. With this scheme, it is possible to simultaneously calibrate the full data set using far fewer compute agents than the number of frequencies at which data are available. We give simulation results to show the feasibility of the proposed multiplexing scheme in simultaneously calibrating a full data set when a limited number of compute agents are available.

  11. Gamma counter calibration system

    International Nuclear Information System (INIS)

    1977-01-01

    A method and apparatus are described for the calibration of a gamma radiation measurement instrument to be used over any of a number of different absolute energy ranges. The method includes the steps of adjusting the overall signal gain associated with pulses which are derived from detected gamma rays, until the instrument is calibrated for a particular absolute energy range; then storing parameter settings corresponding to the adjusted overall signal gain, and repeating the process for other desired absolute energy ranges. The stored settings can be subsequently retrieved and reapplied so that test measurements can be made using a selected one of the absolute energy ranges. Means are provided for adjusting the overall signal gain and a specific technique is disclosed for making coarse, then fine adjustments to the signal gain, for rapid convergence of the required calibration settings. (C.F.)

  12. The Use of Color Sensors for Spectrographic Calibration

    Science.gov (United States)

    Thomas, Neil B.

    2018-04-01

    The wavelength calibration of spectrographs is an essential but challenging task in many disciplines. Calibration is traditionally accomplished by imaging the spectrum of a light source containing features that are known to appear at certain wavelengths and mapping them to their location on the sensor. This is typically required in conjunction with each scientific observation to account for mechanical and optical variations of the instrument over time, which may span years for certain projects. The method presented here investigates the usage of color itself instead of spectral features to calibrate a spectrograph. The primary advantage of such a calibration is that any broad-spectrum light source such as the sky or an incandescent bulb is suitable. This method allows for calibration using the full optical pathway of the instrument instead of incorporating separate calibration equipment that may introduce errors. This paper focuses on the potential for color calibration in the field of radial velocity astronomy, in which instruments must be finely calibrated for long periods of time to detect tiny Doppler wavelength shifts. This method is not restricted to radial velocity, however, and may find application in any field requiring calibrated spectrometers such as sea water analysis, cellular biology, chemistry, atmospheric studies, and so on. This paper demonstrates that color sensors have the potential to provide calibration with greatly reduced complexity.

  13. Rotary mode system initial instrument calibration

    International Nuclear Information System (INIS)

    Johns, B.R.

    1994-01-01

    The attached report contains the vendor calibration procedures used for the initial instrument calibration of the rotary core sampling equipment. The procedures are from approved vendor information files

  14. Post flight analysis of NASA standard star trackers recovered from the solar maximum mission

    Science.gov (United States)

    Newman, P.

    1985-01-01

    The flight hardware returned after the Solar Maximum Mission Repair Mission was analyzed to determine the effects of 4 years in space. The NASA Standard Star Tracker would be a good candidate for such analysis because it is moderately complex and had a very elaborate calibration during the acceptance procedure. However, the recovery process extensively damaged the cathode of the image dissector detector making proper operation of the tracker and a comparison with preflight characteristics impossible. Otherwise, the tracker functioned nominally during testing.

  15. New method to calibrate a spinner anemometer

    DEFF Research Database (Denmark)

    Demurtas, Giorgio; Friis Pedersen, Troels

    2014-01-01

    The spinner anemometer is a wind sensor, based on three one dimensional sonic sensor probes, mounted on the wind turbine spinner, and an algorithm to convert the wind speeds measured by the three sonic sensors to horizontal wind speed, yaw misalignment and flow inclination angle. The conversion...... algorithm utilizes two constants k1 and k2 that are specific to the spinner and blade root design and to the mounting position of the sonic sensors on the spinner. The two constants are calibrated by means of two different test and instrument set-ups. Both calibrations consider the rotor of the wind turbine...... to be stopped during calibration in order for the rotor induction not to influence on the calibration, so that the spinner anemometer measures ”free” wind values in stopped condition. The calibration of flow angle measurements is made by calibration of the ratio of the two algorithm constants k2=k1 = k...

  16. SU-E-T-470: Importance of HU-Mass Density Calibration Technique in Proton Pencil Beam Dose Calculation

    Energy Technology Data Exchange (ETDEWEB)

    Penfold, S; Miller, A [University of Adelaide, Adelaide, SA (Australia)

    2015-06-15

    Purpose: Stoichiometric calibration of Hounsfield Units (HUs) for conversion to proton relative stopping powers (RStPs) is vital for accurate dose calculation in proton therapy. However proton dose distributions are not only dependent on RStP, but also on relative scattering power (RScP) of patient tissues. RScP is approximated from material density but a stoichiometric calibration of HU-density tables is commonly neglected. The purpose of this work was to quantify the difference in calculated dose of a commercial TPS when using HU-density tables based on tissue substitute materials and stoichiometric calibrated ICRU tissues. Methods: Two HU-density calibration tables were generated based on scans of the CIRS electron density phantom. The first table was based directly on measured HU and manufacturer quoted density of tissue substitute materials. The second was based on the same CT scan of the CIRS phantom followed by a stoichiometric calibration of ICRU44 tissue materials. The research version of Pinnacle{sup 3} proton therapy was used to compute dose in a patient CT data set utilizing both HU-density tables. Results: The two HU-density tables showed significant differences for bone tissues; the difference increasing with increasing HU. Differences in density calibration table translated to a difference in calculated RScP of −2.5% for ICRU skeletal muscle and 9.2% for ICRU femur. Dose-volume histogram analysis of a parallel opposed proton therapy prostate plan showed that the difference in calculated dose was negligible when using the two different HU-density calibration tables. Conclusion: The impact of HU-density calibration technique on proton therapy dose calculation was assessed. While differences were found in the calculated RScP of bony tissues, the difference in dose distribution for realistic treatment scenarios was found to be insignificant.

  17. Transient calibration of a groundwater-flow model of Chimacum Creek Basin and vicinity, Jefferson County, Washington: a supplement to Scientific Investigations Report 2013-5160

    Science.gov (United States)

    Jones, Joseph L.; Johnson, Kenneth H.

    2013-01-01

    A steady-state groundwater-flow model described in Scientific Investigations Report 2013-5160, ”Numerical Simulation of the Groundwater-Flow System in Chimacum Creek Basin and Vicinity, Jefferson County, Washington” was developed to evaluate potential future impacts of growth and of water-management strategies on water resources in the Chimacum Creek Basin. This supplement to that report describes the unsuccessful attempt to perform a calibration to transient conditions on the model. The modeled area is about 64 square miles on the Olympic Peninsula in northeastern Jefferson County, Washington. The geologic setting for the model area is that of unconsolidated deposits of glacial and interglacial origin typical of the Puget Sound Lowlands. The hydrogeologic units representing aquifers are Upper Aquifer (UA, roughly corresponding to recessional outwash) and Lower Aquifer (LA, roughly corresponding to advance outwash). Recharge from precipitation is the dominant source of water to the aquifer system; discharge is primarily to marine waters below sea level and to Chimacum Creek and its tributaries. The model is comprised of a grid of 245 columns and 313 rows; cells are a uniform 200 feet per side. There are six model layers, each representing one hydrogeologic unit: (1) Upper Confining unit (UC); (2) Upper Aquifer unit (UA); (3) Middle Confining unit (MC); (4) Lower Aquifer unit (LA); (5) Lower Confining unit (LC); and (6) Bedrock unit (OE). The transient simulation period (October 1994–September 2009) was divided into 180 monthly stress periods to represent temporal variations in recharge, discharge, and storage. An attempt to calibrate the model to transient conditions was unsuccessful due to instabilities stemming from oscillations in groundwater discharge to and recharge from streamflow in Chimacum Creek. The model as calibrated to transient conditions has mean residuals and standard errors of 0.06 ft ±0.45 feet for groundwater levels and 0.48 ± 0.06 cubic

  18. Electron Density Calibration for Radiotherapy Treatment Planning

    International Nuclear Information System (INIS)

    Herrera-Martinez, F.; Rodriguez-Villafuerte, M.; Martinez-Davalos, A.; Ruiz-Trejo, C.; Celis-Lopez, M. A.; Larraga-Gutierrez, J. M.; Garcia-Garduno, A.

    2006-01-01

    Computed tomography (CT) images are used as basic input data for most modern radiosurgery treatment planning systems (TPS). CT data not only provide anatomic information to delineate target volumes, but also allow the introduction of corrections for tissue inhomogeneities into dose calculations during the treatment planning procedure. These corrections involve the determination of a relationship between tissue electron density (ρe) and their corresponding Hounsfield Units (HU). In this work, an elemental analysis of different commercial tissue equivalent materials using Scanning Electron Microscopy was carried out to characterize their chemical composition. The tissue equivalent materials were chosen to ensure a large range of ρe to be included in the CT scanner calibration. A phantom was designed and constructed with these materials to simulate the size of a human head

  19. Sloan Digital Sky Survey Photometric Calibration Revisited

    International Nuclear Information System (INIS)

    Marriner, John

    2012-01-01

    The Sloan Digital Sky Survey calibration is revisited to obtain the most accurate photometric calibration. A small but significant error is found in the flat-fielding of the Photometric telescope used for calibration. Two SDSS star catalogs are compared and the average difference in magnitude as a function of right ascension and declination exhibits small systematic errors in relative calibration. The photometric transformation from the SDSS Photometric Telescope to the 2.5 m telescope is recomputed and compared to synthetic magnitudes computed from measured filter bandpasses.

  20. Sloan Digital Sky Survey Photometric Calibration Revisited

    Energy Technology Data Exchange (ETDEWEB)

    Marriner, John; /Fermilab

    2012-06-29

    The Sloan Digital Sky Survey calibration is revisited to obtain the most accurate photometric calibration. A small but significant error is found in the flat-fielding of the Photometric telescope used for calibration. Two SDSS star catalogs are compared and the average difference in magnitude as a function of right ascension and declination exhibits small systematic errors in relative calibration. The photometric transformation from the SDSS Photometric Telescope to the 2.5 m telescope is recomputed and compared to synthetic magnitudes computed from measured filter bandpasses.

  1. Calibration of EFOSC2 Broadband Linear Imaging Polarimetry

    Science.gov (United States)

    Wiersema, K.; Higgins, A. B.; Covino, S.; Starling, R. L. C.

    2018-03-01

    The European Southern Observatory Faint Object Spectrograph and Camera v2 is one of the workhorse instruments on ESO's New Technology Telescope, and is one of the most popular instruments at La Silla observatory. It is mounted at a Nasmyth focus, and therefore exhibits strong, wavelength and pointing-direction-dependent instrumental polarisation. In this document, we describe our efforts to calibrate the broadband imaging polarimetry mode, and provide a calibration for broadband B, V, and R filters to a level that satisfies most use cases (i.e. polarimetric calibration uncertainty 0.1%). We make our calibration codes public. This calibration effort can be used to enhance the yield of future polarimetric programmes with the European Southern Observatory Faint Object Spectrograph and Camera v2, by allowing good calibration with a greatly reduced number of standard star observations. Similarly, our calibration model can be combined with archival calibration observations to post-process data taken in past years, to form the European Southern Observatory Faint Object Spectrograph and Camera v2 legacy archive with substantial scientific potential.

  2. Proposal of a postal system for Ir-192 sources calibration used in high dose rate brachytherapy with LiF:Mn:Ti thermoluminescent dosemeters

    International Nuclear Information System (INIS)

    Vieira, W.S.; Borges, J.C.; Almeida, C.E.V.

    1998-01-01

    A proposal in order to improve the brachytherapy quality control and to allow postal intercomparison of Ir-192 sources used in high dose rate brachytherapy has been presented. The LiF: Mn: Ti (TLD 100) detector has been selected for such purpose. The experimental array and the TLDs irradiation and calibration techniques, at the treatment units, have been specified in the light of more recent methodology of Ir-192 calibration sources. (Author)

  3. Requirements for the accreditation of a calibration laboratory

    International Nuclear Information System (INIS)

    Palacios, T.A.; Peretti, M.M.

    1993-01-01

    CNEA's activity in calibration is recent but it has a significant development. To assure high quality results, activity must be sustained and improved from day to day. The calibrations laboratory was accredited before Laboratories Qualification Committee, thus adding reliability to its results and making it more competitive when compared to other laboratories not accredited. Among other services given are supervision and follow up of calibrations in laboratories, participation in interlaboratory assays together with other calibration laboratories and assessments on calibration aspects of measuring equipment. (author)

  4. Calibration procedure for Slocum glider deployed optical instruments.

    Science.gov (United States)

    Cetinić, Ivona; Toro-Farmer, Gerardo; Ragan, Matthew; Oberg, Carl; Jones, Burton H

    2009-08-31

    Recent developments in the field of the autonomous underwater vehicles allow the wide usage of these platforms as part of scientific experiments, monitoring campaigns and more. The vehicles are often equipped with sensors measuring temperature, conductivity, chlorophyll a fluorescence (Chl a), colored dissolved organic matter (CDOM) fluorescence, phycoerithrin (PE) fluorescence and spectral volume scattering function at 117 degrees, providing users with high resolution, real time data. However, calibration of these instruments can be problematic. Most in situ calibrations are performed by deploying complementary instrument packages or water samplers in the proximity of the glider. Laboratory calibrations of the mounted sensors are difficult due to the placement of the instruments within the body of the vehicle. For the laboratory calibrations of the Slocum glider instruments we developed a small calibration chamber where we can perform precise calibrations of the optical instruments aboard our glider, as well as sensors from other deployment platforms. These procedures enable us to obtain pre- and post-deployment calibrations for optical fluorescence instruments, which may differ due to the biofouling and other physical damage that can occur during long-term glider deployments. We found that biofouling caused significant changes in the calibration scaling factors of fluorescent sensors, suggesting the need for consistent and repetitive calibrations for gliders as proposed in this paper.

  5. On the calibration of radiotherapy dosemeters in Australia

    International Nuclear Information System (INIS)

    Huntley, R.; Kotler, L.; Webb, D.

    2000-01-01

    Full text: Dosemeters for external beam radiotherapy are calibrated in Australia by ARPANSA, against the national primary standards of exposure and absorbed dose. The primary standards are free air chambers for exposure at low and medium energy X-rays, a graphite cavity chamber for exposure at 60 Co, and a graphite calorimeter for absorbed dose at 60 Co and high energy (MV) X -rays. Radiotherapy dosemeters are calibrated against these standards using a well documented formalism to provide calibration factors suitable for use with dosimetry protocols. A dosemeter usually comprises an ionization chamber connected to an independent electrometer. These are calibrated separately if possible. A combined calibration factor is reported together with the electrometer calibration factor (sensitivity). The dosimetry protocol used in radiotherapy centres in Australia and New Zealand is currently the simplified version of the IAEA TRS277 protocol, published by the New Zealand NRL and recommended by the ACPSEM. This protocol requires the use of an exposure or air kerma calibration factor at 60 Co (Nx or Nk) to evaluate the absorbed dose to air calibration factor N D . The chamber is then placed in a water phantom with its centre displaced from the reference point by p eff . ARPANSA can also supply calibration factors in absorbed dose to water (N D,w ), as required as input to the new IAEA CoP. If an absorbed dose to water calibration factor is used by the radiotherapy centre, the chamber should be placed with its centre at the reference point in the water phantom. ARPANSA has for some years coordinated the participation of Australian radiotherapy centres in the IAEA TLD Quality Audit service. Note that this service does not represent a calibration and should not be referred to as such. The only calibration is that provided by ARPANSA for a reference dosemeter at each radiotherapy centre. As soon as the ANSTO SSDL is operational, calibrations of reference dosemeters will also be

  6. A flow test for calibrating 177 core tubes of 1/5-scale reactor flow model for Yonggwang nuclear units 3 and 4

    International Nuclear Information System (INIS)

    Lee, Byung Jin; Jang, Ho Cheol; Cheong, Jong Sik; Kuh, Jung Eui

    1990-01-01

    A flow test was performed to find out the hydraulic characteristics of every one of 177 core tubes, representing a fuel assembly respectively, as a preparatory step of 1/5 scale reactor flow model test for Yonggwang Nuclear Units (hereafter YGN) 3 and 4. The axial hydraulic resistance of the fuel assembly was simulated in the square core tube with six orifice plates positioned along the tube length; core support structure below each fuel assembly was done in the core upstream geometry section of the test loop. For each core tube the pressure differentials across the inlet, exit orifice plate and overall tube length were measured, along with the flow rates and temperatures of the test fluid. The measured pressure drops were converted to pressure loss or flow metering coefficients. The metering coefficient of the inlet orifice plate was sensitive to the configuration and location of the upstream geometry. The hydraulic resistance of the core tubes were reasonably coincided with a target value and consistent. The polynomial curve fits of the calibrated coefficients for the 177 core tubes were obtained with reasonable data scatters

  7. Radiocarbon calibration - past, present and future

    International Nuclear Information System (INIS)

    Plicht, J. van der

    2004-01-01

    Calibration of the Radiocarbon timescale is traditionally based on tree-rings dated by dendrochronology. At present, the tree-ring curve dates back to about 9900 BC. Beyond this limit, marine datasets extend the present calibration curve INTCAL98 to about 15 600 years ago. Since 1998, a wealth of AMS measurements became available, covering the complete 14 C dating range. No calibration curve can presently be recommended for the older part of the dating range until discrepancies are resolved

  8. Calibration strategies for the Cherenkov Telescope Array

    Science.gov (United States)

    Gaug, Markus; Berge, David; Daniel, Michael; Doro, Michele; Förster, Andreas; Hofmann, Werner; Maccarone, Maria C.; Parsons, Dan; de los Reyes Lopez, Raquel; van Eldik, Christopher

    2014-08-01

    The Central Calibration Facilities workpackage of the Cherenkov Telescope Array (CTA) observatory for very high energy gamma ray astronomy defines the overall calibration strategy of the array, develops dedicated hardware and software for the overall array calibration and coordinates the calibration efforts of the different telescopes. The latter include LED-based light pulsers, and various methods and instruments to achieve a calibration of the overall optical throughput. On the array level, methods for the inter-telescope calibration and the absolute calibration of the entire observatory are being developed. Additionally, the atmosphere above the telescopes, used as a calorimeter, will be monitored constantly with state-of-the-art instruments to obtain a full molecular and aerosol profile up to the stratosphere. The aim is to provide a maximal uncertainty of 10% on the reconstructed energy-scale, obtained through various independent methods. Different types of LIDAR in combination with all-sky-cameras will provide the observatory with an online, intelligent scheduling system, which, if the sky is partially covered by clouds, gives preference to sources observable under good atmospheric conditions. Wide-field optical telescopes and Raman Lidars will provide online information about the height-resolved atmospheric extinction, throughout the field-of-view of the cameras, allowing for the correction of the reconstructed energy of each gamma-ray event. The aim is to maximize the duty cycle of the observatory, in terms of usable data, while reducing the dead time introduced by calibration activities to an absolute minimum.

  9. Upgrading the Medical Physics Calibration Laboratory Towards ISO/IEC 17025: Radiation Standards and Calibration in Diagnostic Radiology

    International Nuclear Information System (INIS)

    Asmaliza Hashim; Muhammad Jamal Md Isa; Abd Aziz Mhd Ramli; Wan Hazlinda Ismail; Norhayati Abdullah; Shahrul Azlan Azizan; Siti Sara Deraman; Nor Azlin Azraai; Md Khairusalih Md Zin

    2010-01-01

    Calibration of quality control (QC) test tools used in diagnostic radiology is legally required under the Ministry of Health (MOH) requirement. The Medical Physics Calibration Laboratory of the Malaysian Nuclear Agency is the national focal point for the calibration of quality control test tools used in diagnostic radiology. The Medical Physics Calibration Laboratory has measurement traceability to primary standard dosimetry laboratory (Physikalisch-Technische Bundesanstalt (PTB)), thus providing an interface between the primary standard dosimetry laboratory and Malaysian hospitals, clinics and license class H holder. The Medical Physics Calibration Laboratory facility is comprised of a constant potential x-ray system with a capability of 160 kV tube and a series of reference and working standard ion chambers. The stability of reference and working standard ion chambers was measured using strontium-90. Dosimetric instruments used in diagnostic radiology is calibrated in terms of air kerma to comply with an International Code of Practices of dosimetry for example IAEA's Technical Report Series number 457. The new series of standard radiation qualities was established based on ISO/IEC 61267. The measurement of beam homogeneity was measured using film and ion chamber to define the field size at certain distance and kV output was measured using the spectrometer and non-invasive kVp meter. The uncertainties measurement was determined with expended uncertainties to a level of confidence of approximately 95% (coverage factor k=2). This paper describes the available facility and the effort of the Medical Physics Calibration Laboratory to upgrade the laboratory towards ISO/IEC 17025. (author)

  10. Individual dosimetry and calibration

    International Nuclear Information System (INIS)

    Otto, T.

    1997-01-01

    In 1996, the Dosimetry and Calibration Section was, as in previous years, mainly engaged in routine tasks: the distribution of over 6000 dosimeters (with a total of more than 10,000 films) every two months and the calibration of about 900 fixed and mobile instruments used in the radiation survey sections of RP group. These tasks were, thanks to an experienced team, well mastered. Special efforts had to be made in a number of areas to modernize the service or to keep it in line with new prescriptions. The Individual Dosimetry Service had to assure that CERN's contracting firms comply with the prescriptions in the Radiation Safety Manual (1996) that had been inspired by the Swiss Ordinance of 1994: Companies must file for authorizations with the Swiss Federal Office for Public Health requiring that in every company an 'Expert in Radiation Protection' be nominated and subsequently trained. CERN's Individual Dosimetry Service is accredited by the Swiss Federal Authorities and works closely together with other, similar services on a rigorous quality assurance programme. Within this framework, CERN was mandated to organize this year the annual Swiss 'Intercomparison of Dosimeters'. All ten accredited dosimetry services - among others those of the Paul Scherrer Institute (PSI) in Villigen and of the four Swiss nuclear power stations - sent dosimeters to CERN, where they were irradiated in CERN's calibration facility with precise photon doses. After return to their origin they were processed and evaluated. The results were communicated to CERN and were compared with the originally given doses. A report on the results was subsequently prepared and submitted to the Swiss 'Group of Experts on Personal Dosimetry'. Reference monitors for photon and neutron radiation were brought to standard laboratories to assure the traceability of CERN's calibration service to the fundamental quantities. For photon radiation, a set of ionization chambers was calibrated in the reference field

  11. CryoSat/SIRAL Cal1 Calibration Orbits

    Science.gov (United States)

    Scagliola, Michele; Fornari, Marco; Bouffard, Jerome; Parrinello, Tommaso

    2017-04-01

    The main payload of CryoSat is a Ku band pulsewidth limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for SAR processing. This allows to reach an along track resolution that is significantly improved with respect to traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed. In fact, not only corrections for transfer function, gain and instrument path delay have to be computed (as in previous altimeters), but also corrections for phase (SAR/SARIn) and phase difference between the two receiving chains (SARIN only). Recalling that the CryoSat's orbit has a high inclination of 92° and it is non-sun-synchronous, the temperature of the SIRAL changes continuously along the orbit with a period of about 480 days and it is also function of the ascending/descending passes. By analysis of the CAL1 calibration corrections, it has been verified that the internal path delay and the instrument gain variation measured on the SIRAL are affected by the thermal status of the instrument and as a consequence they are expected to vary along the orbit. In order to gain knowledge on the calibration corrections (i.e. the instrument behavior) as function of latitude and temperature, it has been planned to command a few number of orbits where only CAL1 calibration acquisitions are continuously performed. The analysis of the CAL1 calibration corrections produced along the Calibration orbits can be also useful to verify whether the current calibration plan is able to provide sufficiently accurate corrections for the instrument acquisitions at any latitude. In 2016, the CryoSat/SIRAL Cal1 Calibration Orbits have been commanded two times, a first time the 20th of July 2016 and a second time the 24th of November 2016, and they

  12. Computerized 50 liter volume calibration system

    International Nuclear Information System (INIS)

    Proffitt, T.H.

    1990-01-01

    A system has been designed for the Savannah River Site that will be used to calibrate product shipping containers. For accountability purposes, it is necessary that these containers be calibrated to a very high precision. The Computerized 50 Liter Volume Calibration System (CVCS), which is based on the Ideal Gas Law (IGL), will use reference volumes with precision of no less ±0.03%, and helium to calibrate the containers to have a total error of no greater than ±0.10%. A statistical interpretation of the system has given a theoretical total calculated error of ±0.08%. Tests with the system will be performed once fabrication is complete to experimentally verify the calculated error. Since the total error was calculated using the worst case scenario, the actual error should be significantly less than the calculated value. The computer controlled, totally automated system is traceable to the National Institute of Standards and Technology. The design, calibration procedure, and statistical interpretation of the system will be discussed. 1 ref

  13. Immune Algorithm Complex Method for Transducer Calibration

    Directory of Open Access Journals (Sweden)

    YU Jiangming

    2014-08-01

    Full Text Available As a key link in engineering test tasks, the transducer calibration has significant influence on accuracy and reliability of test results. Because of unknown and complex nonlinear characteristics, conventional method can’t achieve satisfactory accuracy. An Immune algorithm complex modeling approach is proposed, and the simulated studies on the calibration of third multiple output transducers is made respectively by use of the developed complex modeling. The simulated and experimental results show that the Immune algorithm complex modeling approach can improve significantly calibration precision comparison with traditional calibration methods.

  14. Increased Automation in Stereo Camera Calibration Techniques

    Directory of Open Access Journals (Sweden)

    Brandi House

    2006-08-01

    Full Text Available Robotic vision has become a very popular field in recent years due to the numerous promising applications it may enhance. However, errors within the cameras and in their perception of their environment can cause applications in robotics to fail. To help correct these internal and external imperfections, stereo camera calibrations are performed. There are currently many accurate methods of camera calibration available; however, most or all of them are time consuming and labor intensive. This research seeks to automate the most labor intensive aspects of a popular calibration technique developed by Jean-Yves Bouguet. His process requires manual selection of the extreme corners of a checkerboard pattern. The modified process uses embedded LEDs in the checkerboard pattern to act as active fiducials. Images are captured of the checkerboard with the LEDs on and off in rapid succession. The difference of the two images automatically highlights the location of the four extreme corners, and these corner locations take the place of the manual selections. With this modification to the calibration routine, upwards of eighty mouse clicks are eliminated per stereo calibration. Preliminary test results indicate that accuracy is not substantially affected by the modified procedure. Improved automation to camera calibration procedures may finally penetrate the barriers to the use of calibration in practice.

  15. SU-F-T-215: An Investigation Of Multi-Scanner CT Hounsfield Unit Calibration for Pencil Beam Scanning Proton Therapy Using 3D Gamma Analysis

    International Nuclear Information System (INIS)

    Zhang, J; Li, X; Liu, G; Liu, Q; Liang, J; Ding, X

    2016-01-01

    Purpose: We compare and investigate the dosimetric impacts on pencil beam scanning (PBS) proton treatment plans generated with CT calibration curves from four different CT scanners and one averaged ‘global’ CT calibration curve. Methods: The four CT scanners are located at three different hospital locations within the same health system. CT density calibration curves were collected from these scanners using the same CT calibration phantom and acquisition parameters. Mass density to HU value tables were then commissioned in a commercial treatment planning system. Five disease sites were chosen for dosimetric comparisons at brain, lung, head and neck, adrenal, and prostate. Three types of PBS plans were generated at each treatment site using SFUD, IMPT, and robustness optimized IMPT techniques. 3D dose differences were investigated using 3D Gamma analysis. Results: The CT calibration curves for all four scanners display very similar shapes. Large HU differences were observed at both the high HU and low HU regions of the curves. Large dose differences were generally observed at the distal edges of the beams and they are beam angle dependent. Out of the five treatment sites, lung plans exhibits the most overall range uncertainties and prostate plans have the greatest dose discrepancy. There are no significant differences between the SFUD, IMPT, and the RO-IMPT methods. 3D gamma analysis with 3%, 3 mm criteria showed all plans with greater than 95% passing rate. Two of the scanners with close HU values have negligible dose difference except for lung. Conclusion: Our study shows that there are more than 5% dosimetric differences between different CT calibration curves. PBS treatment plans generated with SFUD, IMPT, and the robustness optimized IMPT has similar sensitivity to the CT density uncertainty. More patient data and tighter gamma criteria based on structure location and size will be used for further investigation.

  16. Dynamic–gravimetric preparation of metrologically traceable primary calibration standards for halogenated greenhouse gases

    Directory of Open Access Journals (Sweden)

    M. Guillevic

    2018-06-01

    Full Text Available For many years, the comparability of measurements obtained with various instruments within a global-scale air quality monitoring network has been ensured by anchoring all results to a unique suite of reference gas mixtures, also called a primary calibration scale. Such suites of reference gas mixtures are usually prepared and then stored over decades in pressurised cylinders by a designated laboratory. For the halogenated gases which have been measured over the last 40 years, this anchoring method is highly relevant as measurement reproducibility is currently much better ( <  1 %, k  =  2 or 95 % confidence interval than the expanded uncertainty of a reference gas mixture (usually  >  2 %. Meanwhile, newly emitted halogenated gases are already measured in the atmosphere at pmol mol−1 levels, while still lacking an established reference standard. For compounds prone to adsorption on material surfaces, it is difficult to evaluate mixture stability and thus variations in the molar fractions over time in cylinders at pmol mol−1 levels.To support atmospheric monitoring of halogenated gases, we create new primary calibration scales for SF6 (sulfur hexafluoride, HFC-125 (pentafluoroethane, HFO-1234yf (or HFC-1234yf, 2,3,3,3-tetrafluoroprop-1-ene, HCFC-132b (1,2-dichloro-1,1-difluoroethane and CFC-13 (chlorotrifluoromethane. The preparation method, newly applied to halocarbons, is dynamic and gravimetric: it is based on the permeation principle followed by dynamic dilution and cryo-filling of the mixture in cylinders. The obtained METAS-2017 primary calibration scales are made of 11 cylinders containing these five substances at near-ambient and slightly varying molar fractions. Each prepared molar fraction is traceable to the realisation of SI units (International System of Units and is assigned an uncertainty estimate following international guidelines (JCGM, 2008, ranging from 0.6 % for SF6 to 1.3 % (k

  17. Moving Beyond Streamflow Observations: Lessons From A Multi-Objective Calibration Experiment in the Mississippi Basin

    Science.gov (United States)

    Koppa, A.; Gebremichael, M.; Yeh, W. W. G.

    2017-12-01

    Calibrating hydrologic models in large catchments using a sparse network of streamflow gauges adversely affects the spatial and temporal accuracy of other water balance components which are important for climate-change, land-use and drought studies. This study combines remote sensing data and the concept of Pareto-Optimality to address the following questions: 1) What is the impact of streamflow (SF) calibration on the spatio-temporal accuracy of Evapotranspiration (ET), near-surface Soil Moisture (SM) and Total Water Storage (TWS)? 2) What is the best combination of fluxes that can be used to calibrate complex hydrological models such that both the accuracy of streamflow and the spatio-temporal accuracy of ET, SM and TWS is preserved? The study area is the Mississippi Basin in the United States (encompassing HUC-2 regions 5,6,7,9,10 and 11). 2003 and 2004, two climatologically average years are chosen for calibration and validation of the Noah-MP hydrologic model. Remotely sensed ET data is sourced from GLEAM, SM from ESA-CCI and TWS from GRACE. Single objective calibration is carried out using DDS Algorithm. For Multi objective calibration PA-DDS is used. First, the Noah-MP model is calibrated using a single objective function (Minimize Mean Square Error) for the outflow from the 6 HUC-2 sub-basins for 2003. Spatial correlograms are used to compare the spatial structure of ET, SM and TWS between the model and the remote sensing data. Spatial maps of RMSE and Mean Error are used to quantify the impact of calibrating streamflow on the accuracy of ET, SM and TWS estimates. Next, a multi-objective calibration experiment is setup to determine the pareto optimal parameter sets (pareto front) for the following cases - 1) SF and ET, 2) SF and SM, 3) SF and TWS, 4) SF, ET and SM, 5) SF, ET and TWS, 6) SF, SM and TWS, 7) SF, ET, SM and TWS. The best combination of fluxes that provides the optimal trade-off between accurate streamflow and preserving the spatio

  18. SU-E-T-749: Thorough Calibration of MOSFET Dosimeters

    International Nuclear Information System (INIS)

    Plenkovich, D; Thomas, J

    2015-01-01

    Purpose: To improve the accuracy of the MOSFET calibration procedure by performing the measurement several times and calculating the average value of the calibration factor for various photon and electron energies. Methods: The output of three photon and six electron beams of Varian Trilogy linear accelerator SN 5878 was calibrated. Five reinforced standard sensitivity MOSFET dosimeters were placed in the calibration jig and connected to the Reader Module. As the backscatter material was used 7 cm of Virtual Water. The MOSFET dosimeters were covered with 1.5 cm thick bolus for the regular and SRS 6 MV beams, 3 cm bolus for 15 MV beam, 1.5 cm bolus for 6 MeV electron beam, and 2 cm bolus for the electron energies of 9, 12, 15, 18, and 22 MeV. The dosimeters were exposed to 100 MU, and the calibration factor was determined using the mobileMOSFET software. To improve the accuracy of calibration, this procedure was repeated ten times and the calibration factors were averaged. Results: As the number of calibrations was increasing the variability of calibration factors of different dosimeters was decreasing. After ten calibrations, the calibration factors for all five dosimeters were within 1% of one another for all energies, except 6 MV SRS photons and 6 MeV electrons, for which the variability was 2%. Conclusions: The described process results in calibration factors which are almost independent of modality or energy. Once calibrated, the dosimeters may be used for in-vivo dosimetry or for daily verification of the beam output. Measurement of the radiation dose under bolus and scatter to the eye are examples of frequent use of calibrated MOSFET dosimeters. The calibration factor determined for full build-up is used under these circumstances. To the best of our knowledge, such thorough procedure for calibrating MOSFET dosimeters has not been reported previously. Best Medical Canada provided MOSFET dosimeters for this project

  19. Ionosphere Delay Calibration and Calibration Errors for Satellite Navigation of Aircraft

    Science.gov (United States)

    Harris, Ian; Manucci, Anthony; Iijima, Byron; Lindqwister, Ulf; Muna, Demitri; Pi, Xiaoqing; Wilson, Brian

    2000-01-01

    The Federal Aviation Administration (FAA) is implementing a satellite-based navigation system for aircraft using the Global Positioning System (GPS). Positioning accuracy of a few meters will be achieved by broadcasting corrections to the direct GPS signal. These corrections are derived using the wide-area augmentation system (WAAS), which includes a ground network of at least 24 GPS receivers across the Continental US (CONUS). WAAS will provide real-time total electron content (TEC) measurements that can be mapped to fixed grid points using a real-time mapping algorithm. These TECs will be converted into vertical delay corrections for the GPS L1 frequency and broadcast to users every five minutes via geosynchronous satellite. Users will convert these delays to slant calibrations along their own lines-of-sight (LOS) to GPS satellites. Uncertainties in the delay calibrations will also be broadcast, allowing users to estimate the uncertainty of their position. To maintain user safety without reverting to excessive safety margins an empirical model of user calibration errors has been developed. WAAS performance depends on factors that include geographic location (errors increase near WAAS borders), and ionospheric conditions, such as the enhanced spatial electron density gradients found during ionospheric storms.

  20. Soil Moisture Active/Passive (SMAP) Radiometer Subband Calibration and Calibration Drift

    Science.gov (United States)

    Peng, Jinzheng; Piepmeier, Jeffrey R.; De Amici, Giovanni; Mohammed, Priscilla

    2016-01-01

    The SMAP is one of four first-tier missions recommended by the US National Research Council's Committee on Earth Science and Applications from Space (Earth Science and Applications from Space: National Imperatives for the Next Decade and Beyond, Space Studies Board, National Academies Press, 2007)]. The observatory was launched on Jan 31, 2015. The goal of the SMAP is to measure the global soil moisture and freeze/thaw from space. The L-band radiometer is the passive portion of the spaceborne instrument. It measures all four Stokes antenna temperatures and outputs counts. The Level 1B Brightness Temperature (L1B_TB) science algorithm converts radiometer counts to the Earths surface brightness temperature. The results are reported in the radiometer level 1B data product together with the calibrated antenna temperature (TA) and all of the corrections to the unwanted sources contribution. The calibrated L1B data product are required to satisfy the overall radiometer error budget of 1.3 K needed to meet the soil moisture requirement of 0.04 volumetric fraction uncertainty and the calibration drift requirement of no larger than 0.4 K per month.

  1. Soil Moisture Active Passive (SMAP) Radiometer Subband Calibration and Calibration Drift

    Science.gov (United States)

    Peng, Jinzheng; Piepmeier, Jeffrey R.; De Amici, Giovanni; Mohammed, Priscilla N.

    2016-01-01

    The SMAP is one of four first-tier missions recommended by the US National Research Council's Committee on Earth Science and Applications from Space (Earth Science and Applications from Space: National Imperatives for the Next Decade and Beyond, Space Studies Board, National Academies Press, 2007). The observatory was launched on Jan 31, 2015. The goal of the SMAP is to measure the global soil moisture and freeze/thaw from space. The L-band radiometer is the passive portion of the spaceborne instrument. It measures all four Stokes antenna temperatures and outputs counts. The Level 1B Brightness Temperature (L1B_TB) science algorithm converts radiometer counts to the Earths surface brightness temperature. The results are reported in the radiometer level 1B data product together with the calibrated antenna temperature (TA) and all of the corrections to the unwanted sources contribution. The calibrated L1B data product are required to satisfy the overall radiometer error budget of 1.3 K needed to meet the soil moisture requirement of 0.04 volumetric fraction uncertainty and the calibration drift requirement of no larger than 0.4 K per month.

  2. Radiocarbon calibration - past, present and future

    Energy Technology Data Exchange (ETDEWEB)

    Plicht, J. van der E-mail: plicht@phys.rug.nl

    2004-08-01

    Calibration of the Radiocarbon timescale is traditionally based on tree-rings dated by dendrochronology. At present, the tree-ring curve dates back to about 9900 BC. Beyond this limit, marine datasets extend the present calibration curve INTCAL98 to about 15 600 years ago. Since 1998, a wealth of AMS measurements became available, covering the complete {sup 14}C dating range. No calibration curve can presently be recommended for the older part of the dating range until discrepancies are resolved.

  3. Calibration of alpha surface contamination monitor

    International Nuclear Information System (INIS)

    Freitas, I.S.M. de; Goncalez, O.L.

    1990-01-01

    In this work, the results, as well as the methodology, of the calibration of an alpha surface contamination monitor are presented. The calibration factors are obtained by least-squares fitting with effective variance. (author)

  4. HPS instrument calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Masse, F.X; Eisenhower, E.H.; Swinth, K.L.

    1993-12-31

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory.

  5. The Observability Calibration Test Development Framework

    Energy Technology Data Exchange (ETDEWEB)

    Endicott-Popovsky, Barbara E.; Frincke, Deborah A.

    2007-06-20

    Abstract— Formal standards, precedents, and best practices for verifying and validating the behavior of low layer network devices used for digital evidence-collection on networks are badly needed— initially so that these can be employed directly by device owners and data users to document the behaviors of these devices for courtroom presentation, and ultimately so that calibration testing and calibration regimes are established and standardized as common practice for both vendors and their customers [1]. The ultimate intent is to achieve a state of confidence in device calibration that allows the network data gathered by them to be relied upon by all parties in a court of law. This paper describes a methodology for calibrating forensic-ready low layer network devices based on the Flaw Hypothesis Methodology [2,3].

  6. XRD alignment, calibration and performance

    International Nuclear Information System (INIS)

    Davy, L.

    2002-01-01

    Full text: The quality of any diffractometer system is very much dependent on the alignment, calibration and performance. The three subjects are very much related. Firstly, you must know how to carry out the full diffractometer alignment. XRD alignment is easy once you know how. The presentation will show you step by step to carry out the full alignment. Secondly, you need to know how to calibrate the diffractometer system. The presentation will show you how to calibrate the goniometer, detector etc. Thirdly, to prove the system is working within the manufacturer specification. The presentation will show you how to carry out the resolution, reproducibility and linearity test. Copyright (2002) Australian X-ray Analytical Association Inc

  7. CERN radiation protection (RP) calibration facilities

    Energy Technology Data Exchange (ETDEWEB)

    Pozzi, Fabio

    2016-04-14

    Radiation protection calibration facilities are essential to ensure the correct operation of radiation protection instrumentation. Calibrations are performed in specific radiation fields according to the type of instrument to be calibrated: neutrons, photons, X-rays, beta and alpha particles. Some of the instruments are also tested in mixed radiation fields as often encountered close to high-energy particle accelerators. Moreover, calibration facilities are of great importance to evaluate the performance of prototype detectors; testing and measuring the response of a prototype detector to well-known and -characterized radiation fields contributes to improving and optimizing its design and capabilities. The CERN Radiation Protection group is in charge of performing the regular calibrations of all CERN radiation protection devices; these include operational and passive dosimeters, neutron and photon survey-meters, and fixed radiation detectors to monitor the ambient dose equivalent, H*(10), inside CERN accelerators and at the CERN borders. A new state-of-the-art radiation protection calibration facility was designed, constructed and commissioned following the related ISO recommendations to replace the previous ageing (more than 30 years old) laboratory. In fact, the new laboratory aims also at the official accreditation according to the ISO standards in order to be able to release certified calibrations. Four radiation fields are provided: neutrons, photons and beta sources and an X-ray generator. Its construction did not only involve a pure civil engineering work; many radiation protection studies were performed to provide a facility that could answer the CERN calibration needs and fulfill all related safety requirements. Monte Carlo simulations have been confirmed to be a valuable tool for the optimization of the building design, the radiation protection aspects, e.g. shielding, and, as consequence, the overall cost. After the source and irradiator installation

  8. Radiometric and spectral calibrations of the Geostationary Imaging Fourier Transform Spectrometer (GIFTS) using principle component analysis

    Science.gov (United States)

    Tian, Jialin; Smith, William L.; Gazarik, Michael J.

    2008-10-01

    The ultimate remote sensing benefits of the high resolution Infrared radiance spectrometers will be realized with their geostationary satellite implementation in the form of imaging spectrometers. This will enable dynamic features of the atmosphere's thermodynamic fields and pollutant and greenhouse gas constituents to be observed for revolutionary improvements in weather forecasts and more accurate air quality and climate predictions. As an important step toward realizing this application objective, the Geostationary Imaging Fourier Transform Spectrometer (GIFTS) Engineering Demonstration Unit (EDU) was successfully developed under the NASA New Millennium Program, 2000-2006. The GIFTS-EDU instrument employs three focal plane arrays (FPAs), which gather measurements across the long-wave IR (LWIR), short/mid-wave IR (SMWIR), and visible spectral bands. The raw GIFTS interferogram measurements are radiometrically and spectrally calibrated to produce radiance spectra, which are further processed to obtain atmospheric profiles via retrieval algorithms. The radiometric calibration is achieved using internal blackbody calibration references at ambient (260 K) and hot (286 K) temperatures. The absolute radiometric performance of the instrument is affected by several factors including the FPA off-axis effect, detector/readout electronics induced nonlinearity distortions, and fore-optics offsets. The GIFTS-EDU, being the very first imaging spectrometer to use ultra-high speed electronics to readout its large area format focal plane array detectors, operating at wavelengths as large as 15 microns, possessed non-linearity's not easily removable in the initial calibration process. In this paper, we introduce a refined calibration technique that utilizes Principle Component (PC) analysis to compensate for instrument distortions and artifacts remaining after the initial radiometric calibration process, thus, further enhance the absolute calibration accuracy. This method is

  9. Calibration of laboratory equipment and its intermediate verification

    International Nuclear Information System (INIS)

    Remedi, Jorge O.

    2011-01-01

    When a laboratory wants to prove that he has technical competence to carry out tests or calibrations must demonstrate that it has complied with certain requirements that establish , among others, the mandatory : calibrate or verify equipment before putting it into service in order to ensure that it meets to the specifications of laboratory equipment to keep records evidencing the checks that equipment complies with the specification ; perform intermediate checks for maintain confidence in the calibration status of the equipment , ensure that the operation is checked and calibration status of equipment when the equipment goes outside the direct control of the laboratory , before be returned to service, establish a program and procedure for the calibration of equipment; show how determined the calibration periods of their equipment as well as evidence that intermediate checks are suitable for the calibration periods. However, some confusion is observed as to the meaning of the terms 'calibration' and 'verification' of a computer. This paper analyzes applicable documentation and suggests that the differences are generated in part by translations and by characterization concepts upon its usage, that is, if it is legal metrology or assessment conformity. Therefore, this study aims to characterize both concepts , fundamentals to zoom distinguish , outline appropriate strategies for calibration and verification activities to ensure the compliance with regulatory requirements [es

  10. Some problems in calibrating surface contamination meters

    International Nuclear Information System (INIS)

    Chen Zigen; LI Xingyuan; Shuai Xiaoping.

    1984-01-01

    It is necessary that instruments are calibrated accurately in order to obtain reliable survey data of surface contamination. Some problems in calibrating surface contamination meters are expounded in this paper. Measurement comparison for beta surface contamination meters is organized within limited scope, thus survey quality is understood, questions are discovered, significance of calibration is expounded further. (Author)

  11. CryoSat-2 SIRAL Calibration and Performance

    Science.gov (United States)

    Fornari, M.; Scagliola, M.; Tagliani, N.; Parrinello, T.

    2012-12-01

    The main payload of CryoSat-2 is a Ku band pulse-width limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for azimuth processing. This allows to reach an along track resolution of about 250 meters which is a significant improvement over traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed, including both an internal and external calibration. The internal calibration monitors the instrument impulse response and the transfer function, like traditional altimeters. In addition to that, the interferometer requires a special calibration developed ad hoc for SIRAL. The external calibration is performed with the use of a ground transponder, located in Svalbard, which receives SIRAL signal and sends the echo back to the satellite. Internal calibration data are processed on ground by the CryoSat-2 Instrument Processing Facility (IPF1) and then applied to the science data. In December 2012, two and a half years of calibration data will be available, which will be shown in this poster. The external calibration (transponder) data are processed and analyzed independently from the operational chain. The use of an external transponder has been very useful to determine instrument performance and for the tuning of the on-ground processor. This poster presents the transponder results in terms of range noise and datation error.

  12. Providing primary standard calibrations beyond 20 MHz

    International Nuclear Information System (INIS)

    Bickley, C J; Zeqiri, B; Robinson, S P

    2004-01-01

    The number of applications of medical ultrasound utilising frequencies in excess of 20 MHz has shown a consistent increase over recent years. Coupled with the commercial availability of wide-bandwidth hydrophones whose response extends beyond 40 MHz, this has driven a growing need to develop hydrophone calibration techniques at elevated frequencies. The current National Physical Laboratory primary standard method of calibrating hydrophones is based on an optical interferometer. This has been in operation for around 20 years and provides traceability over the frequency range of 0.3 to 20 MHz. More recently, calibrations carried out using the interferometer have been extended to 60 MHz, although the uncertainties associated with these calibrations are poor, being in excess of ±20% at high frequencies. Major contributions to the degraded calibration uncertainties arise from poor signal-to-noise at higher frequencies, the frequency response of the photodiodes used and the noise floor of the instrument. To improve the uncertainty of hydrophone calibrations above 20 MHz, it has been necessary to build and commission a new interferometer. Important features of the new primary standard are its use of a higher power laser to improve the signal-to-noise ratio, along with photodiodes whose greater bandwidth to improve the overall frequency response. This paper describes the design of key aspects of the new interferometer. It also presents some initial results of the performance assessment, including a detailed comparison of calibrations of NPL reference membrane hydrophones, undertaken using old and new interferometers for calibration up to 40 MHz

  13. Automated intraoperative calibration for prostate cancer brachytherapy

    International Nuclear Information System (INIS)

    Kuiran Chen, Thomas; Heffter, Tamas; Lasso, Andras; Pinter, Csaba; Abolmaesumi, Purang; Burdette, E. Clif; Fichtinger, Gabor

    2011-01-01

    Purpose: Prostate cancer brachytherapy relies on an accurate spatial registration between the implant needles and the TRUS image, called ''calibration''. The authors propose a new device and a fast, automatic method to calibrate the brachytherapy system in the operating room, with instant error feedback. Methods: A device was CAD-designed and precision-engineered, which mechanically couples a calibration phantom with an exact replica of the standard brachytherapy template. From real-time TRUS images acquired from the calibration device and processed by the calibration system, the coordinate transformation between the brachytherapy template and the TRUS images was computed automatically. The system instantly generated a report of the target reconstruction accuracy based on the current calibration outcome. Results: Four types of validation tests were conducted. First, 50 independent, real-time calibration trials yielded an average of 0.57 ± 0.13 mm line reconstruction error (LRE) relative to ground truth. Second, the averaged LRE was 0.37 ± 0.25 mm relative to ground truth in tests with six different commercial TRUS scanners operating at similar imaging settings. Furthermore, testing with five different commercial stepper systems yielded an average of 0.29 ± 0.16 mm LRE relative to ground truth. Finally, the system achieved an average of 0.56 ± 0.27 mm target registration error (TRE) relative to ground truth in needle insertion tests through the template in a water tank. Conclusions: The proposed automatic, intraoperative calibration system for prostate cancer brachytherapy has achieved high accuracy, precision, and robustness.

  14. Supervised Vicarious Calibration (SVC of Multi-Source Hyperspectral Remote-Sensing Data

    Directory of Open Access Journals (Sweden)

    Anna Brook

    2015-05-01

    Full Text Available Introduced in 2011, the supervised vicarious calibration (SVC approach is a promising approach to radiometric calibration and atmospheric correction of airborne hyperspectral (HRS data. This paper presents a comprehensive study by which the SVC method has been systematically examined and a complete protocol for its practical execution has been established—along with possible limitations encountered during the campaign. The technique was applied to multi-sourced HRS data in order to: (1 verify the at-sensor radiometric calibration and (2 obtain radiometric and atmospheric correction coefficients. Spanning two select study sites along the southeast coast of France, data were collected simultaneously by three airborne sensors (AisaDUAL, AHS and CASI-1500i aboard two aircrafts (CASA of National Institute for Aerospace Technology INTA ES and DORNIER 228 of NERC-ARSF Centre UK. The SVC ground calibration site was assembled along sand dunes near Montpellier and the thematic data were acquired from other areas in the south of France (Salon-de-Provence, Marseille, Avignon and Montpellier on 28 October 2010 between 12:00 and 16:00 UTC. The results of this study confirm that the SVC method enables reliable inspection and, if necessary, in-situ fine radiometric recalibration of airborne hyperspectral data. Independent of sensor or platform quality, the SVC approach allows users to improve at-sensor data to obtain more accurate physical units and subsequently improved reflectance information. Flight direction was found to be important, whereas the flight altitude posed very low impact. The numerous rules and major outcomes of this experiment enable a new standard of atmospherically corrected data based on better radiometric output. Future research should examine the potential of SVC to be applied to super-and-hyperspectral data obtained from on-orbit sensors.

  15. Quality assurance programs at the PNL calibrations laboratory

    International Nuclear Information System (INIS)

    Piper, R.K.; McDonald, J.C.; Fox, R.A.; Eichner, F.N.

    1993-03-01

    The calibrations laboratory at Pacific Northwest Laboratory (PNL) serves as a radiological standardization facility for personnel and environmental dosimetry and radiological survey instruments. As part of this function, the calibrations laboratory must maintain radiological reference fields with calibrations traceable to the National Institute of Standards and Technology (NIST). This task is accomplished by a combination of (1) sources or reference instruments calibrated at or by NIST, (2) measurement quality assurance (MQA) interactions with NIST, and (3) rigorous internal annual and quarterly calibration verifications. This paper describes a representative sample of the facilities, sources, and actions used to maintain accurate and traceable fields

  16. Calibration of the SNO+ experiment

    Science.gov (United States)

    Maneira, J.; Falk, E.; Leming, E.; Peeters, S.; SNO+ Collaboration.

    2017-09-01

    The main goal of the SNO+ experiment is to perform a low-background and high-isotope-mass search for neutrinoless double-beta decay, employing 780 tonnes of liquid scintillator loaded with tellurium, in its initial phase at 0.5% by mass for a total mass of 1330 kg of 130Te. The SNO+ physics program includes also measurements of geo- and reactor neutrinos, supernova and solar neutrinos. Calibrations are an essential component of the SNO+ data-taking and analysis plan. The achievement of the physics goals requires both an extensive and regular calibration. This serves several goals: the measurement of several detector parameters, the validation of the simulation model and the constraint of systematic uncertainties on the reconstruction and particle identification algorithms. SNO+ faces stringent radiopurity requirements which, in turn, largely determine the materials selection, sealing and overall design of both the sources and deployment systems. In fact, to avoid frequent access to the inner volume of the detector, several permanent optical calibration systems have been developed and installed outside that volume. At the same time, the calibration source internal deployment system was re-designed as a fully sealed system, with more stringent material selection, but following the same working principle as the system used in SNO. This poster described the overall SNO+ calibration strategy, discussed the several new and innovative sources, both optical and radioactive, and covered the developments on source deployment systems.

  17. Energy calibration of a multilayer photon detector

    International Nuclear Information System (INIS)

    Johnson, R.A.

    1983-01-01

    The job of energy calibration was broken into three parts: gain normalization of all equivalent elements; determination of the functions for conversion of pulse height to energy; and gain stabilization. It is found that calorimeter experiments are no better than their calibration systems - calibration errors will be the major source of error at high energies. Redundance is found to be necessary - the system should be designed such that every element could be replaced during the life of the experiment. It is found to be important to have enough data taken during calibration runs and during the experiment to be able to sort out where the calibration problems were after the experiment is over. Each layer was normalized independently with electrons, and then the pulse height to energy conversion was determined with photons. The primary method of gain stabilization used the light flasher system

  18. Traceable calibration of hospital 192Ir HDR sources

    International Nuclear Information System (INIS)

    Govinda Rajan, K.N.; Bhatt, B.C.; Pendse, A.M.; Kannan, V.

    2002-01-01

    Presently, no primary standard exists for the standardization of remote afterloading 192 Ir HDR sources. These sources are, therefore, being standardized by a few Secondary Standard Dosimetry Laboratories (SSDLs), in terms of Air Kerma Strength (AKS) or Reference Air Kerma Rate (RAKR) using a 0.6 cc Farmer type chamber, set up as an Interim Standard. These SSDLs offer calibration to well type of ionization chambers that are normally used by the hospitals for calibrating the 192 lr HDR source. Presently, in many countries, including India, well chambers are not commercially available. Nor do these countries offer any calibration service for 192 lr HDR source. With the result users make use of well chambers imported from different countries with their calibration traceable to the country of origin. Since no intercomparisons between these countries have been reported, the measurement consistency between hospitals becomes questionable. The problem is compounded by the fact that these chambers are used for several years without re-calibration since no calibration service is locally available. For instance, in India, the chambers have been in use in hospitals, since 1994, without a second calibration. Not all hospitals use the well chamber for the calibration of the 192 lr HDR source. Many hospitals make use of 0.6 cc chambers, in air, at short source to chamber distances, for measuring the AKS of the source. The latter method is prone to much larger inaccuracy due to the use of very short source to chamber distances without proper calibration jigs, use of 60 Co calibration factor for 192 Ir HDR source calibrations, neglecting correction factors for room scatter, fluence non-uniformity, use of arbitrary buildup factors for the buildup cap of the chamber etc. A comparison of the procedures used at hospitals revealed that various arbitrary methods are in use at hospitals. An indigenously developed well chamber was calibrated against a Reference Standard traceable to the

  19. Real-time continuous glucose monitoring shows high accuracy within 6 hours after sensor calibration: a prospective study.

    Directory of Open Access Journals (Sweden)

    Xiao-Yan Yue

    Full Text Available Accurate and timely glucose monitoring is essential in intensive care units. Real-time continuous glucose monitoring system (CGMS has been advocated for many years to improve glycemic management in critically ill patients. In order to determine the effect of calibration time on the accuracy of CGMS, real-time subcutaneous CGMS was used in 18 critically ill patients. CGMS sensor was calibrated with blood glucose measurements by blood gas/glucose analyzer every 12 hours. Venous blood was sampled every 2 to 4 hours, and glucose concentration was measured by standard central laboratory device (CLD and by blood gas/glucose analyzer. With CLD measurement as reference, relative absolute difference (mean±SD in CGMS and blood gas/glucose analyzer were 14.4%±12.2% and 6.5%±6.2%, respectively. The percentage of matched points in Clarke error grid zone A was 74.8% in CGMS, and 98.4% in blood gas/glucose analyzer. The relative absolute difference of CGMS obtained within 6 hours after sensor calibration (8.8%±7.2% was significantly less than that between 6 to 12 hours after calibration (20.1%±13.5%, p<0.0001. The percentage of matched points in Clarke error grid zone A was also significantly higher in data sets within 6 hours after calibration (92.4% versus 57.1%, p<0.0001. In conclusion, real-time subcutaneous CGMS is accurate in glucose monitoring in critically ill patients. CGMS sensor should be calibrated less than 6 hours, no matter what time interval recommended by manufacturer.

  20. Crop physiology calibration in the CLM

    Directory of Open Access Journals (Sweden)

    I. Bilionis

    2015-04-01

    scalable and adaptive scheme based on sequential Monte Carlo (SMC. The model showed significant improvement of crop productivity with the new calibrated parameters. We demonstrate that the calibrated parameters are applicable across alternative years and different sites.

  1. Calibration and uncertainty in electromagnetic fields measuring methods

    International Nuclear Information System (INIS)

    Anglesio, L.; Crotti, G.; Borsero, M.; Vizio, G.

    1999-01-01

    Calibration and reliability in electromagnetic field measuring methods are assured by calibration of measuring instruments. In this work are illustrated systems for generation of electromagnetic fields at low and high frequency, calibration standard and accuracy [it

  2. SU-E-T-442: Geometric Calibration and Verification of a GammaPod Breast SBRT System

    Energy Technology Data Exchange (ETDEWEB)

    Yu, C [Univ Maryland School of Medicine, Baltimore, MD (United States); Xcision Medical Systems, Columbia, MD (United States); Niu, Y; Maton, P; Hoban, P [Xcision Medical Systems, Columbia, MD (United States); Mutaf, Y [Univ Maryland School of Medicine, Baltimore, MD (United States)

    2015-06-15

    Purpose: The first GammaPod™ unit for prone stereotactic treatment of early stage breast cancer has recently been installed and calibrated. Thirty-six rotating circular Co-60 beams focus dose at an isocenter that traverses throughout a breast target via continuous motion of the treatment table. The breast is immobilized and localized using a vacuum-assisted stereotactic cup system that is fixed to the table during treatment. Here we report on system calibration and on verification of geometric and dosimetric accuracy. Methods: Spatial calibration involves setting the origin of each table translational axis within the treatment control system such that the relationship between beam isocenter and table geometry is consistent with that assumed by the treatment planning system. A polyethylene QA breast phantom inserted into an aperture in the patient couch is used for calibration and verification. The comparison is performed via fiducial-based registration of measured single-isocenter dose profiles (radiochromic film) with kernel dose profiles. With the table calibrations applied, measured relative dose distributions were compared with TPS calculations for single-isocenter and dynamic (many-isocenter) treatment plans. Further, table motion accuracy and linearity was tested via comparison of planned control points with independent encoder readouts. Results: After table calibration, comparison of measured and calculated single-isocenter dose profiles show agreement to within 0.5 mm for each axis. Gamma analysis of measured vs calculated profiles with 3%/2mm criteria yields a passing rate of >99% and >98% for single-isocenter and dynamic plans respectively. This also validates the relative dose distributions produced by the TPS. Measured table motion accuracy was within 0.05 mm for all translational axes. Conclusion: GammaPod table coordinate calibration is a straightforward process that yields very good agreement between planned and measured relative dose distributions

  3. High accuracy injection circuit for the calibration of a large pixel sensor matrix

    International Nuclear Information System (INIS)

    Quartieri, E.; Comotti, D.; Manghisoni, M.

    2013-01-01

    Semiconductor pixel detectors, for particle tracking and vertexing in high energy physics experiments as well as for X-ray imaging, in particular for synchrotron light sources and XFELs, require a large area sensor matrix. This work will discuss the design and the characterization of a high-linearity, low dispersion injection circuit to be used for pixel-level calibration of detector readout electronics in a large pixel sensor matrix. The circuit provides a useful tool for the characterization of the readout electronics of the pixel cell unit for both monolithic active pixel sensors and hybrid pixel detectors. In the latter case, the circuit allows for precise analogue test of the readout channel already at the chip level, when no sensor is connected. Moreover, it provides a simple means for calibration of readout electronics once the detector has been connected to the chip. Two injection techniques can be provided by the circuit: one for a charge sensitive amplification and the other for a transresistance readout channel. The aim of the paper is to describe the architecture and the design guidelines of the calibration circuit, which has been implemented in a 130 nm CMOS technology. Moreover, experimental results of the proposed injection circuit will be presented in terms of linearity and dispersion

  4. Comparison of infusion pumps calibration methods

    Science.gov (United States)

    Batista, Elsa; Godinho, Isabel; do Céu Ferreira, Maria; Furtado, Andreia; Lucas, Peter; Silva, Claudia

    2017-12-01

    Nowadays, several types of infusion pump are commonly used for drug delivery, such as syringe pumps and peristaltic pumps. These instruments present different measuring features and capacities according to their use and therapeutic application. In order to ensure the metrological traceability of these flow and volume measuring equipment, it is necessary to use suitable calibration methods and standards. Two different calibration methods can be used to determine the flow error of infusion pumps. One is the gravimetric method, considered as a primary method, commonly used by National Metrology Institutes. The other calibration method, a secondary method, relies on an infusion device analyser (IDA) and is typically used by hospital maintenance offices. The suitability of the IDA calibration method was assessed by testing several infusion instruments at different flow rates using the gravimetric method. In addition, a measurement comparison between Portuguese Accredited Laboratories and hospital maintenance offices was performed under the coordination of the Portuguese Institute for Quality, the National Metrology Institute. The obtained results were directly related to the used calibration method and are presented in this paper. This work has been developed in the framework of the EURAMET projects EMRP MeDD and EMPIR 15SIP03.

  5. Calibration of a Modified Californium Shuffler

    International Nuclear Information System (INIS)

    Sadowski, E.T.; Armstrong, F.; Oldham, R.; Ceo, R.; Williams, N.

    1995-01-01

    A californium shuffler originally designed to assay hollow cylindrical pieces of UA1 has been modified to assay solid cylinders. Calibration standards were characterized via chemical analysis of the molten UA1 taken during casting of the standards. The melt samples yielded much more reliable characterization data than drill samples taken from standards after the standards had solidified. By normalizing one well-characterized calibration curve to several standards at different enrichments, a relatively small number of standards was required to develop an enrichment-dependent calibration. The precision of this shuffler is 0.65%, and the typical random and systematic uncertainties are 0.53% and 0.73%, respectively, for a six minute assay of an ingot containing approximately 700 grams of 235 U. This paper will discuss (1) the discrepancies encountered when UA1 standards were characterized via melt samples versus drill samples, (2) a calibration methodology employing a small number of standards, and (3) a comparison of results from a previously unused shuffler with an existing shuffler. A small number of UA1 standards have been characterized using samples from the homogeneous molten state and have yielded enrichment-dependent and enrichment-independent calibration curves on two different shufflers

  6. ASD FieldSpec Calibration Setup and Techniques

    Science.gov (United States)

    Olive, Dan

    2001-01-01

    This paper describes the Analytical Spectral Devices (ASD) Fieldspec Calibration Setup and Techniques. The topics include: 1) ASD Fieldspec FR Spectroradiometer; 2) Components of Calibration; 3) Equipment list; 4) Spectral Setup; 5) Spectral Calibration; 6) Radiometric and Linearity Setup; 7) Radiometric setup; 8) Datadets Required; 9) Data files; and 10) Field of View Measurement. This paper is in viewgraph form.

  7. Calibration of Ground-based Lidar instrument

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Gómez Arranz, Paula

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements. The comparison of the lidar measurements of the wind direction with that from wind vanes...

  8. Preliminary report on NTS spectral gamma logging and calibration models

    International Nuclear Information System (INIS)

    Mathews, M.A.; Warren, R.G.; Garcia, S.R.; Lavelle, M.J.

    1985-01-01

    Facilities are now available at the Nevada Test Site (NTS) in Building 2201 to calibrate spectral gamma logging equipment in environments of low radioactivity. Such environments are routinely encountered during logging of holes at the NTS. Four calibration models were delivered to Building 2201 in January 1985. Each model, or test pit, consists of a stone block with a 12-inch diameter cored borehole. Preliminary radioelement values from the core for the test pits range from 0.58 to 3.83% potassium (K), 0.48 to 29.11 ppm thorium (Th), and 0.62 to 40.42 ppm uranium (U). Two satellite holes, U19ab number2 and U19ab number3, were logged during the winter of 1984-1985. The response of these logs correlates with contents of the naturally radioactive elements K. Th. and U determined in samples from petrologic zones that occur within these holes. Based on these comparisons, the spectral gamma log aids in the recognition and mapping of subsurface stratigraphic units and alteration features associated with unusual concentration of these radioactive elements, such as clay-rich zones

  9. Radiation protection dosimetry and calibrations

    International Nuclear Information System (INIS)

    Verhavere, Ph.

    2007-01-01

    At the SCK-CEN different specialised services are delivered for a whole range of external and internal customers in the radiation protection area. For the expertise group of radiation protection dosimetry and calibrations, these services are organized in four different laboratories: dosimetry, anthropogammametry, nuclear calibrations and non-nuclear calibrations. The services are given by a dedicated technical staff who has experience in the handling of routine and specialised cases. The scientific research that is performed by the expertise group makes sure that state-of-the-art techniques are being used, and that constant improvements and developments are implemented. Quality Assurance is an important aspect for the different services, and accreditation according national and international standards is achieved for all laboratories

  10. Calibrating a novel multi-sensor physical activity measurement system

    International Nuclear Information System (INIS)

    John, D; Sasaki, J E; Howe, C A; Freedson, P S; Liu, S; Gao, R X; Staudenmayer, J

    2011-01-01

    Advancing the field of physical activity (PA) monitoring requires the development of innovative multi-sensor measurement systems that are feasible in the free-living environment. The use of novel analytical techniques to combine and process these multiple sensor signals is equally important. This paper describes a novel multi-sensor 'integrated PA measurement system' (IMS), the lab-based methodology used to calibrate the IMS, techniques used to predict multiple variables from the sensor signals, and proposes design changes to improve the feasibility of deploying the IMS in the free-living environment. The IMS consists of hip and wrist acceleration sensors, two piezoelectric respiration sensors on the torso, and an ultraviolet radiation sensor to obtain contextual information (indoors versus outdoors) of PA. During lab-based calibration of the IMS, data were collected on participants performing a PA routine consisting of seven different ambulatory and free-living activities while wearing a portable metabolic unit (criterion measure) and the IMS. Data analyses on the first 50 adult participants are presented. These analyses were used to determine if the IMS can be used to predict the variables of interest. Finally, physical modifications for the IMS that could enhance the feasibility of free-living use are proposed and refinement of the prediction techniques is discussed

  11. The Chandra Source Catalog 2.0: Calibrations

    Science.gov (United States)

    Graessle, Dale E.; Evans, Ian N.; Rots, Arnold H.; Allen, Christopher E.; Anderson, Craig S.; Budynkiewicz, Jamie A.; Burke, Douglas; Chen, Judy C.; Civano, Francesca Maria; D'Abrusco, Raffaele; Doe, Stephen M.; Evans, Janet D.; Fabbiano, Giuseppina; Gibbs, Danny G., II; Glotfelty, Kenny J.; Grier, John D.; Hain, Roger; Hall, Diane M.; Harbo, Peter N.; Houck, John C.; Lauer, Jennifer L.; Laurino, Omar; Lee, Nicholas P.; Martínez-Galarza, Juan Rafael; McCollough, Michael L.; McDowell, Jonathan C.; Miller, Joseph; McLaughlin, Warren; Morgan, Douglas L.; Mossman, Amy E.; Nguyen, Dan T.; Nichols, Joy S.; Nowak, Michael A.; Paxson, Charles; Plummer, David A.; Primini, Francis Anthony; Siemiginowska, Aneta; Sundheim, Beth A.; Tibbetts, Michael; Van Stone, David W.; Zografou, Panagoula

    2018-01-01

    Among the many enhancements implemented for the release of Chandra Source Catalog (CSC) 2.0 are improvements in the processing calibration database (CalDB). We have included a thorough overhaul of the CalDB software used in the processing. The software system upgrade, called "CalDB version 4," allows for a more rational and consistent specification of flight configurations and calibration boundary conditions. Numerous improvements in the specific calibrations applied have also been added. Chandra's radiometric and detector response calibrations vary considerably with time, detector operating temperature, and position on the detector. The CalDB has been enhanced to provide the best calibrations possible to each observation over the fifteen-year period included in CSC 2.0. Calibration updates include an improved ACIS contamination model, as well as updated time-varying gain (i.e., photon energy) and quantum efficiency maps for ACIS and HRC-I. Additionally, improved corrections for the ACIS quantum efficiency losses due to CCD charge transfer inefficiency (CTI) have been added for each of the ten ACIS detectors. These CTI corrections are now time and temperature-dependent, allowing ACIS to maintain a 0.3% energy calibration accuracy over the 0.5-7.0 keV range for any ACIS source in the catalog. Radiometric calibration (effective area) accuracy is estimated at ~4% over that range. We include a few examples where improvements in the Chandra CalDB allow for improved data reduction and modeling for the new CSC.This work has been supported by NASA under contract NAS 8-03060 to the Smithsonian Astrophysical Observatory for operation of the Chandra X-ray Center.

  12. Radiation energy calibrating system and method

    International Nuclear Information System (INIS)

    Jacobson, A.F.

    1980-01-01

    A radiation energy calibrating system and method which uses a pair of calibrated detectors for measurements of radiation intensity from x-ray tubes for a non-invasive determination of the electrical characteristics; I.E., the tube potential and/or current

  13. Systems and methods of eye tracking calibration

    DEFF Research Database (Denmark)

    2014-01-01

    Methods and systems to facilitate eye tracking control calibration are provided. One or more objects are displayed on a display of a device, where the one or more objects are associated with a function unrelated to a calculation of one or more calibration parameters. The one or more calibration...... parameters relate to a calibration of a calculation of gaze information of a user of the device, where the gaze information indicates where the user is looking. While the one or more objects are displayed, eye movement information associated with the user is determined, which indicates eye movement of one...... or more eye features associated with at least one eye of the user. The eye movement information is associated with a first object location of the one or more objects. The one or more calibration parameters are calculated based on the first object location being associated with the eye movement information....

  14. Direct illumination LED calibration for telescope photometry

    International Nuclear Information System (INIS)

    Barrelet, E.; Juramy, C.

    2008-01-01

    A calibration method for telescope photometry, based on the direct illumination of a telescope with a calibrated light source regrouping multiple LEDs, is proposed. Its purpose is to calibrate the instrument response. The main emphasis of the proposed method is the traceability of the calibration process and a continuous monitoring of the instrument in order to maintain a 0.2% accuracy over a period of years. Its specificity is to map finely the response of the telescope and its camera as a function of all light ray parameters. This feature is essential to implement a computer model of the instrument representing the variation of the overall light collection efficiency of each pixel for various filter configurations. We report on hardware developments done for SNDICE, the first application of this direct illumination calibration system which will be installed in Canada France Hawaii telescope (CFHT) for its leading supernova experiment (SNLS)

  15. Indoor calibration for stereoscopic camera STC: a new method

    Science.gov (United States)

    Simioni, E.; Re, C.; Da Deppo, V.; Naletto, G.; Borrelli, D.; Dami, M.; Ficai Veltroni, I.; Cremonese, G.

    2017-11-01

    In the framework of the ESA-JAXA BepiColombo mission to Mercury, the global mapping of the planet will be performed by the on-board Stereo Camera (STC), part of the SIMBIO-SYS suite [1]. In this paper we propose a new technique for the validation of the 3D reconstruction of planetary surface from images acquired with a stereo camera. STC will provide a three-dimensional reconstruction of Mercury surface. The generation of a DTM of the observed features is based on the processing of the acquired images and on the knowledge of the intrinsic and extrinsic parameters of the optical system. The new stereo concept developed for STC needs a pre-flight verification of the actual capabilities to obtain elevation information from stereo couples: for this, a stereo validation setup to get an indoor reproduction of the flight observing condition of the instrument would give a much greater confidence to the developed instrument design. STC is the first stereo satellite camera with two optical channels converging in a unique sensor. Its optical model is based on a brand new concept to minimize mass and volume and to allow push-frame imaging. This model imposed to define a new calibration pipeline to test the reconstruction method in a controlled ambient. An ad-hoc indoor set-up has been realized for validating the instrument designed to operate in deep space, i.e. in-flight STC will have to deal with source/target essentially placed at infinity. This auxiliary indoor setup permits on one side to rescale the stereo reconstruction problem from the operative distance in-flight of 400 km to almost 1 meter in lab; on the other side it allows to replicate different viewing angles for the considered targets. Neglecting for sake of simplicity the Mercury curvature, the STC observing geometry of the same portion of the planet surface at periherm corresponds to a rotation of the spacecraft (SC) around the observed target by twice the 20° separation of each channel with respect to nadir

  16. High-dose secondary calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Humphreys, J.C. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program.

  17. Calibration services for medical applications of radiation

    Energy Technology Data Exchange (ETDEWEB)

    DeWerd, L.A.

    1993-12-31

    Calibration services for the medical community applications of radiation involve measuring radiation precisely and having traceability to the National Institute of Standards and Technology (NIST). Radiation therapy applications involve the use of ionization chambers and electrometers for external beams and well-type ionization chamber systems as well as radioactive sources for brachytherapy. Diagnostic x-ray applications involve ionization chamber systems and devices to measure other parameters of the x-ray machine, such as non-invasive kVp meters. Calibration laboratories have been established to provide radiation calibration services while maintaining traceability to NIST. New radiation applications of the medical community spur investigation to provide the future calibration needs.

  18. Calibration services for medical applications of radiation

    International Nuclear Information System (INIS)

    DeWerd, L.A.

    1993-01-01

    Calibration services for the medical community applications of radiation involve measuring radiation precisely and having traceability to the National Institute of Standards and Technology (NIST). Radiation therapy applications involve the use of ionization chambers and electrometers for external beams and well-type ionization chamber systems as well as radioactive sources for brachytherapy. Diagnostic x-ray applications involve ionization chamber systems and devices to measure other parameters of the x-ray machine, such as non-invasive kVp meters. Calibration laboratories have been established to provide radiation calibration services while maintaining traceability to NIST. New radiation applications of the medical community spur investigation to provide the future calibration needs

  19. High-dose secondary calibration laboratory accreditation program

    International Nuclear Information System (INIS)

    Humphreys, J.C.

    1993-01-01

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program

  20. Photometric Calibration of Consumer Video Cameras

    Science.gov (United States)

    Suggs, Robert; Swift, Wesley, Jr.

    2007-01-01

    Equipment and techniques have been developed to implement a method of photometric calibration of consumer video cameras for imaging of objects that are sufficiently narrow or sufficiently distant to be optically equivalent to point or line sources. Heretofore, it has been difficult to calibrate consumer video cameras, especially in cases of image saturation, because they exhibit nonlinear responses with dynamic ranges much smaller than those of scientific-grade video cameras. The present method not only takes this difficulty in stride but also makes it possible to extend effective dynamic ranges to several powers of ten beyond saturation levels. The method will likely be primarily useful in astronomical photometry. There are also potential commercial applications in medical and industrial imaging of point or line sources in the presence of saturation.This development was prompted by the need to measure brightnesses of debris in amateur video images of the breakup of the Space Shuttle Columbia. The purpose of these measurements is to use the brightness values to estimate relative masses of debris objects. In most of the images, the brightness of the main body of Columbia was found to exceed the dynamic ranges of the cameras. A similar problem arose a few years ago in the analysis of video images of Leonid meteors. The present method is a refined version of the calibration method developed to solve the Leonid calibration problem. In this method, one performs an endto- end calibration of the entire imaging system, including not only the imaging optics and imaging photodetector array but also analog tape recording and playback equipment (if used) and any frame grabber or other analog-to-digital converter (if used). To automatically incorporate the effects of nonlinearity and any other distortions into the calibration, the calibration images are processed in precisely the same manner as are the images of meteors, space-shuttle debris, or other objects that one seeks to

  1. SU-E-T-391: Evaluation of Image Parameters Impact On the CT Calibration Curve for Proton Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Xiao, Z; Reyhan, M; Huang, Q; Zhang, M; Yue, N; Chen, T [Rutgers University, New Brunswick, NJ (United States)

    2015-06-15

    Purpose: The calibration of the Hounsfield units (HU) to relative proton stopping powers (RSP) is a crucial component in assuring the accurate delivery of proton therapy dose distributions to patients. The purpose of this work is to assess the uncertainty of CT calibration considering the impact of CT slice thickness, position of the plug within the phantom and phantom sizes. Methods: Stoichiometric calibration method was employed to develop the CT calibration curve. Gammex 467 tissue characterization phantom was scanned in Tomotherapy Cheese phantom and Gammex 451 phantom by using a GE CT scanner. Each plug was individually inserted into the same position of inner and outer ring of phantoms at each time, respectively. 1.25 mm and 2.5 mm slice thickness were used. Other parameters were same. Results: HU of selected human tissues were calculated based on fitted coefficient (Kph, Kcoh and KKN), and RSP were calculated according to the Bethe-Bloch equation. The calibration curve was obtained by fitting cheese phantom data with 1.25 mm thickness. There is no significant difference if the slice thickness, phantom size, position of plug changed in soft tissue. For boney structure, RSP increases up to 1% if the phantom size and the position of plug changed but keep the slice thickness the same. However, if the slice thickness varied from the one in the calibration curve, 0.5%–3% deviation would be expected depending on the plug position. The Inner position shows the obvious deviation (averagely about 2.5%). Conclusion: RSP shows a clinical insignificant deviation in soft tissue region. Special attention may be required when using a different slice thickness from the calibration curve for boney structure. It is clinically practical to address 3% deviation due to different thickness in the definition of clinical margins.

  2. 1987 calibration of the TFTR neutron spectrometers

    International Nuclear Information System (INIS)

    Barnes, C.W.; Strachan, J.D.; Princeton Univ., NJ

    1989-12-01

    The 3 He neutron spectrometer used for measuring ion temperatures and the NE213 proton recoil spectrometer used for triton burnup measurements were absolutely calibrated with DT and DD neutron generators placed inside the TFTR vacuum vessel. The details of the detector response and calibration are presented. Comparisons are made to the neutron source strengths measured from other calibrated systems. 23 refs., 19 figs., 6 tabs

  3. SRHA calibration curve

    Data.gov (United States)

    U.S. Environmental Protection Agency — an UV calibration curve for SRHA quantitation. This dataset is associated with the following publication: Chang, X., and D. Bouchard. Surfactant-Wrapped Multiwalled...

  4. Cumulative sum quality control for calibrated breast density measurements

    International Nuclear Information System (INIS)

    Heine, John J.; Cao Ke; Beam, Craig

    2009-01-01

    Purpose: Breast density is a significant breast cancer risk factor. Although various methods are used to estimate breast density, there is no standard measurement for this important factor. The authors are developing a breast density standardization method for use in full field digital mammography (FFDM). The approach calibrates for interpatient acquisition technique differences. The calibration produces a normalized breast density pixel value scale. The method relies on first generating a baseline (BL) calibration dataset, which required extensive phantom imaging. Standardizing prospective mammograms with calibration data generated in the past could introduce unanticipated error in the standardized output if the calibration dataset is no longer valid. Methods: Sample points from the BL calibration dataset were imaged approximately biweekly over an extended timeframe. These serial samples were used to evaluate the BL dataset reproducibility and quantify the serial calibration accuracy. The cumulative sum (Cusum) quality control method was used to evaluate the serial sampling. Results: There is considerable drift in the serial sample points from the BL calibration dataset that is x-ray beam dependent. Systematic deviation from the BL dataset caused significant calibration errors. This system drift was not captured with routine system quality control measures. Cusum analysis indicated that the drift is a sign of system wear and eventual x-ray tube failure. Conclusions: The BL calibration dataset must be monitored and periodically updated, when necessary, to account for sustained system variations to maintain the calibration accuracy.

  5. Cumulative sum quality control for calibrated breast density measurements

    Energy Technology Data Exchange (ETDEWEB)

    Heine, John J.; Cao Ke; Beam, Craig [Cancer Prevention and Control Division, Moffitt Cancer Center, 12902 Magnolia Drive, Tampa, Florida 33612 (United States); Division of Epidemiology and Biostatistics, School of Public Health, University of Illinois at Chicago, 1603 W. Taylor St., Chicago, Illinois 60612 (United States)

    2009-12-15

    Purpose: Breast density is a significant breast cancer risk factor. Although various methods are used to estimate breast density, there is no standard measurement for this important factor. The authors are developing a breast density standardization method for use in full field digital mammography (FFDM). The approach calibrates for interpatient acquisition technique differences. The calibration produces a normalized breast density pixel value scale. The method relies on first generating a baseline (BL) calibration dataset, which required extensive phantom imaging. Standardizing prospective mammograms with calibration data generated in the past could introduce unanticipated error in the standardized output if the calibration dataset is no longer valid. Methods: Sample points from the BL calibration dataset were imaged approximately biweekly over an extended timeframe. These serial samples were used to evaluate the BL dataset reproducibility and quantify the serial calibration accuracy. The cumulative sum (Cusum) quality control method was used to evaluate the serial sampling. Results: There is considerable drift in the serial sample points from the BL calibration dataset that is x-ray beam dependent. Systematic deviation from the BL dataset caused significant calibration errors. This system drift was not captured with routine system quality control measures. Cusum analysis indicated that the drift is a sign of system wear and eventual x-ray tube failure. Conclusions: The BL calibration dataset must be monitored and periodically updated, when necessary, to account for sustained system variations to maintain the calibration accuracy.

  6. Field calibration and modification of scs design equation for predicting length of border under local conditions

    International Nuclear Information System (INIS)

    Choudhary, M.R.; Mustafa, U.S.

    2009-01-01

    Field tests were conducted to calibrate the existing SCS design equation in determining field border length using field data of different field lengths during 2nd and 3rd irrigations under local conditions. A single ring infiltrometer was used to estimate the water movement into and through the irrigated soil profile and in estimating the coefficients of Kostiakov infiltration function. Measurements of the unit discharge and time of advance were carried out during different irrigations on wheat irrigated fields having clay loam soil. The collected field data were used to calibrate the existing SCS design equation developed by USDA for testing its validity under local field conditions. SCS equation was modified further to improve its applicability. Results from the study revealed that the Kostiakov model over predicted the coefficients, which in turn overestimated the water advance length for boarder in the selected field using existing SCS design equation. However, the calibrated SCS design equation after parametric modification produced more satisfactory results encouraging the scientists to make its use at larger scale. (author)

  7. Comparison of Calibration Techniques for Low-Cost Air Quality Monitoring

    Science.gov (United States)

    Malings, C.; Ramachandran, S.; Tanzer, R.; Kumar, S. P. N.; Hauryliuk, A.; Zimmerman, N.; Presto, A. A.

    2017-12-01

    Assessing the intra-city spatial distribution and temporal variability of air quality can be facilitated by a dense network of monitoring stations. However, the cost of implementing such a network can be prohibitive if high-quality but high-cost monitoring systems are used. To this end, the Real-time Affordable Multi-Pollutant (RAMP) sensor package has been developed at the Center for Atmospheric Particle Studies of Carnegie Mellon University, in collaboration with SenSevere LLC. This self-contained unit can measure up to five gases out of CO, SO2, NO, NO2, O3, VOCs, and CO2, along with temperature and relative humidity. Responses of individual gas sensors can vary greatly even when exposed to the same ambient conditions. Those of VOC sensors in particular were observed to vary by a factor-of-8, which suggests that each sensor requires its own calibration model. To this end, we apply and compare two different calibration methods to data collected by RAMP sensors collocated with a reference monitor station. The first method, random forest (RF) modeling, is a rule-based method which maps sensor responses to pollutant concentrations by implementing a trained sequence of decision rules. RF modeling has previously been used for other RAMP gas sensors by the group, and has produced precise calibrated measurements. However, RF models can only predict pollutant concentrations within the range observed in the training data collected during the collocation period. The second method, Gaussian process (GP) modeling, is a probabilistic Bayesian technique whereby broad prior estimates of pollutant concentrations are updated using sensor responses to generate more refined posterior predictions, as well as allowing predictions beyond the range of the training data. The accuracy and precision of these techniques are assessed and compared on VOC data collected during the summer of 2017 in Pittsburgh, PA. By combining pollutant data gathered by each RAMP sensor and applying

  8. The Geostationary Lightning Mapper: Its Performance and Calibration

    Science.gov (United States)

    Christian, H. J., Jr.

    2015-12-01

    The Geostationary Lightning Mapper (GLM) has been developed to be an operational instrument on the GOES-R series of spacecraft. The GLM is a unique instrument, unlike other meteorological instruments, both in how it operates and in the information content that it provides. Instrumentally, it is an event detector, rather than an imager. While processing almost a billion pixels per second with 14 bits of resolution, the event detection process reduces the required telemetry bandwidth by almost 105, thus keeping the telemetry requirements modest and enabling efficient ground processing that leads to rapid data distribution to operational users. The GLM was designed to detect about 90 percent of the total lightning flashes within its almost hemispherical field of view. Based on laboratory calibration, we expect the on-orbit detection efficiency to be closer to 85%, making it the highest performing, large area coverage total lightning detector. It has a number of unique design features that will enable it have near uniform special resolution over most of its field of view and to operate with minimal impact on performance during solar eclipses. The GLM has no dedicated on-orbit calibration system, thus the ground-based calibration provides the bases for the predicted radiometric performance. A number of problems were encountered during the calibration of Flight Model 1. The issues arouse from GLM design features including its wide field of view, fast lens, the narrow-band interference filters located in both object and collimated space and the fact that the GLM is inherently a event detector yet the calibration procedures required both calibration of images and events. The GLM calibration techniques were based on those developed for the Lightning Imaging Sensor calibration, but there are enough differences between the sensors that the initial GLM calibration suggested that it is significantly more sensitive than its design parameters. The calibration discrepancies have

  9. 21 CFR 882.1925 - Ultrasonic scanner calibration test block.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Ultrasonic scanner calibration test block. 882... Ultrasonic scanner calibration test block. (a) Identification. An ultrasonic scanner calibration test block is a block of material with known properties used to calibrate ultrasonic scanning devices (e.g., the...

  10. Absolute calibration of SARAL/AltiKa in Kavaratti during its initial calibration-validation phase

    Digital Repository Service at National Institute of Oceanography (India)

    Babu, K.N.; Shukla, A.K.; Suchandra, A.B.; ArunKumar, S.V.V.; Bonnefond, P.; Testut, L.; Mehra, P.; Laurain, O.

    globally distributed region will offer assessment of the altimetry system, and allow us to check in specific conditions leading to different estimation of absolute bias of the instrument (Shum et al. 2003). In collaboration with National Institute... of Oceanography (NIO), Goa, Space Applica- tions Centre–Indian Space Research Organisation (SAC-ISRO) established a calibration- verification site in Kavaratti. This site offers a number of advantages as a calibration site for altimeters. Having very small land...

  11. Instrumentation calibration

    International Nuclear Information System (INIS)

    Mack, D.A.

    1976-08-01

    Procedures for the calibration of different types of laboratory equipment are described. Provisions for maintaining the integrity of reference and working standards traceable back to a national standard are discussed. Methods of validation and certification methods are included. An appendix lists available publications and services of national standardizing agencies

  12. Review of Calibration Methods for Scheimpflug Camera

    Directory of Open Access Journals (Sweden)

    Cong Sun

    2018-01-01

    Full Text Available The Scheimpflug camera offers a wide range of applications in the field of typical close-range photogrammetry, particle image velocity, and digital image correlation due to the fact that the depth-of-view of Scheimpflug camera can be greatly extended according to the Scheimpflug condition. Yet, the conventional calibration methods are not applicable in this case because the assumptions used by classical calibration methodologies are not valid anymore for cameras undergoing Scheimpflug condition. Therefore, various methods have been investigated to solve the problem over the last few years. However, no comprehensive review exists that provides an insight into recent calibration methods of Scheimpflug cameras. This paper presents a survey of recent calibration methods of Scheimpflug cameras with perspective lens, including the general nonparametric imaging model, and analyzes in detail the advantages and drawbacks of the mainstream calibration models with respect to each other. Real data experiments including calibrations, reconstructions, and measurements are performed to assess the performance of the models. The results reveal that the accuracies of the RMM, PLVM, PCIM, and GNIM are basically equal, while the accuracy of GNIM is slightly lower compared with the other three parametric models. Moreover, the experimental results reveal that the parameters of the tangential distortion are likely coupled with the tilt angle of the sensor in Scheimpflug calibration models. The work of this paper lays the foundation of further research of Scheimpflug cameras.

  13. Actuator-Assisted Calibration of Freehand 3D Ultrasound System.

    Science.gov (United States)

    Koo, Terry K; Silvia, Nathaniel

    2018-01-01

    Freehand three-dimensional (3D) ultrasound has been used independently of other technologies to analyze complex geometries or registered with other imaging modalities to aid surgical and radiotherapy planning. A fundamental requirement for all freehand 3D ultrasound systems is probe calibration. The purpose of this study was to develop an actuator-assisted approach to facilitate freehand 3D ultrasound calibration using point-based phantoms. We modified the mathematical formulation of the calibration problem to eliminate the need of imaging the point targets at different viewing angles and developed an actuator-assisted approach/setup to facilitate quick and consistent collection of point targets spanning the entire image field of view. The actuator-assisted approach was applied to a commonly used cross wire phantom as well as two custom-made point-based phantoms (original and modified), each containing 7 collinear point targets, and compared the results with the traditional freehand cross wire phantom calibration in terms of calibration reproducibility, point reconstruction precision, point reconstruction accuracy, distance reconstruction accuracy, and data acquisition time. Results demonstrated that the actuator-assisted single cross wire phantom calibration significantly improved the calibration reproducibility and offered similar point reconstruction precision, point reconstruction accuracy, distance reconstruction accuracy, and data acquisition time with respect to the freehand cross wire phantom calibration. On the other hand, the actuator-assisted modified "collinear point target" phantom calibration offered similar precision and accuracy when compared to the freehand cross wire phantom calibration, but it reduced the data acquisition time by 57%. It appears that both actuator-assisted cross wire phantom and modified collinear point target phantom calibration approaches are viable options for freehand 3D ultrasound calibration.

  14. The 2007 ESO Instrument Calibration Workshop

    CERN Document Server

    Kaufer, Andreas; ESO Workshop

    2008-01-01

    The 2007 ESO Instrument Calibration workshop brought together more than 120 participants with the objective to a) foster the sharing of information, experience and techniques between observers, instrument developers and instrument operation teams, b) review the actual precision and limitations of the applied instrument calibration plans, and c) collect the current and future requirements by the ESO users. These present proceedings include the majority of the workshop’s contributions and document the status quo of instrument calibration at ESO in large detail. Topics covered are: Optical Spectro-Imagers, Optical Multi-Object Spectrographs, NIR and MIR Spectro-Imagers, High-Resolution Spectrographs, Integral Field Spectrographs, Adaptive Optics Instruments, Polarimetric Instruments, Wide Field Imagers, Interferometric Instruments as well as other crucial aspects such as data flow, quality control, data reduction software and atmospheric effects. It was stated in the workshop that "calibration is a life-long l...

  15. A calibration method for phoswich detectors

    CERN Document Server

    Gawlikowicz, W; Schröder, W U

    2002-01-01

    A new method of particle identification and energy calibration is proposed for phoswich detectors consisting of a fast plastic scintillator optically coupled to a CsI(Tl) crystal. In this method, functional relationships between integrated yields of three properly selected portions of the observed composite light output signal are parameterized in terms of three physical components (overlapping in time) of this signal. Then, a complete set of calibration constants is derived, linking these integrated partial yields to particle atomic numbers (Z) and energies (E). The calibration is based on the correlations between partial yields observed in actual data from the physics or 'production' runs, and energy deposits calculated for detector elements. The method is found suitable for a wide range of incident particle species and energies, largely eliminating the need for costly and time-consuming conventional calibration experiments with a variety of beams, targets, and bombarding energies.

  16. Temperature corrected-calibration of GRACE's accelerometer

    Science.gov (United States)

    Encarnacao, J.; Save, H.; Siemes, C.; Doornbos, E.; Tapley, B. D.

    2017-12-01

    Since April 2011, the thermal control of the accelerometers on board the GRACE satellites has been turned off. The time series of along-track bias clearly show a drastic change in the behaviour of this parameter, while the calibration model has remained unchanged throughout the entire mission lifetime. In an effort to improve the quality of the gravity field models produced at CSR in future mission-long re-processing of GRACE data, we quantify the added value of different calibration strategies. In one approach, the temperature effects that distort the raw accelerometer measurements collected without thermal control are corrected considering the housekeeping temperature readings. In this way, one single calibration strategy can be consistently applied during the whole mission lifetime, since it is valid to thermal the conditions before and after April 2011. Finally, we illustrate that the resulting calibrated accelerations are suitable for neutral thermospheric density studies.

  17. Carbon resistor pressure gauge calibration at low stresses

    International Nuclear Information System (INIS)

    Cunningham, Bruce; Vandersall, Kevin S.; Niles, Angela M.; Greenwood, Daniel W.; Garcia, Frank; Forbes, Jerry W.; Wilson, William H.

    2002-01-01

    The 470 Ohm carbon resistor gauge has been used in the stress range up to 4-5 GPa for highly heterogeneous materials and/or divergent flow experiments. The attractiveness of the gauge is its rugged nature, simple construction, low cost, reproducibility, and survivability in dynamic events. Gauge drawbacks are the long time response to pressure equilibration and gauge resistance hysteresis. In the regime below 0.4 GPa, gauge calibration has been extrapolated. Because of the need for calibration data within this low stress regime, calibration experiments were performed using a split-Hopkinson bar, drop tower apparatus, and gas pressure chamber. Since the performance of the gauge at elevated temperatures is a concern, the change in resistance due to heating at atmospheric pressure was also investigated. Details of the various calibration arrangements and the results are discussed and compared to a calibration curve fit to previously published calibration data

  18. Nintendo WII remotes provide a reconfigurable tool-changing unit with an automatic calibration capability

    Directory of Open Access Journals (Sweden)

    Collins, James

    2014-08-01

    Full Text Available Modular machines within the reconfigurable manufacturing paradigm require auxiliary modules to enhance the system’s capability. A tool-changing unit was developed as one of these auxiliary modules. The unit had to be able to adapt itself efficiently to changes in the configuration of the machine it was servicing. This necessitated the development of a real- time 3D tracking system in order for the unit to sense alterations in the position of the spindle to which it was delivering tools. An economic positioning system was produced using Nintendo Wii remotes. This paper presents the development, implementation, and testing of this positioning system.

  19. CPAS Preflight Drop Test Analysis Process

    Science.gov (United States)

    Englert, Megan E.; Bledsoe, Kristin J.; Romero, Leah M.

    2015-01-01

    Throughout the Capsule Parachute Assembly System (CPAS) drop test program, the CPAS Analysis Team has developed a simulation and analysis process to support drop test planning and execution. This process includes multiple phases focused on developing test simulations and communicating results to all groups involved in the drop test. CPAS Engineering Development Unit (EDU) series drop test planning begins with the development of a basic operational concept for each test. Trajectory simulation tools include the Flight Analysis and Simulation Tool (FAST) for single bodies, and the Automatic Dynamic Analysis of Mechanical Systems (ADAMS) simulation for the mated vehicle. Results are communicated to the team at the Test Configuration Review (TCR) and Test Readiness Review (TRR), as well as at Analysis Integrated Product Team (IPT) meetings in earlier and intermediate phases of the pre-test planning. The ability to plan and communicate efficiently with rapidly changing objectives and tight schedule constraints is a necessity for safe and successful drop tests.

  20. Radiometric inter-sensor cross-calibration uncertainty using a traceable high accuracy reference hyperspectral imager

    Science.gov (United States)

    Gorroño, Javier; Banks, Andrew C.; Fox, Nigel P.; Underwood, Craig

    2017-08-01

    Optical earth observation (EO) satellite sensors generally suffer from drifts and biases relative to their pre-launch calibration, caused by launch and/or time in the space environment. This places a severe limitation on the fundamental reliability and accuracy that can be assigned to satellite derived information, and is particularly critical for long time base studies for climate change and enabling interoperability and Analysis Ready Data. The proposed TRUTHS (Traceable Radiometry Underpinning Terrestrial and Helio-Studies) mission is explicitly designed to address this issue through re-calibrating itself directly to a primary standard of the international system of units (SI) in-orbit and then through the extension of this SI-traceability to other sensors through in-flight cross-calibration using a selection of Committee on Earth Observation Satellites (CEOS) recommended test sites. Where the characteristics of the sensor under test allows, this will result in a significant improvement in accuracy. This paper describes a set of tools, algorithms and methodologies that have been developed and used in order to estimate the radiometric uncertainty achievable for an indicative target sensor through in-flight cross-calibration using a well-calibrated hyperspectral SI-traceable reference sensor with observational characteristics such as TRUTHS. In this study, Multi-Spectral Imager (MSI) of Sentinel-2 and Landsat-8 Operational Land Imager (OLI) is evaluated as an example, however the analysis is readily translatable to larger-footprint sensors such as Sentinel-3 Ocean and Land Colour Instrument (OLCI) and Visible Infrared Imaging Radiometer Suite (VIIRS). This study considers the criticality of the instrumental and observational characteristics on pixel level reflectance factors, within a defined spatial region of interest (ROI) within the target site. It quantifies the main uncertainty contributors in the spectral, spatial, and temporal domains. The resultant tool

  1. Calibration of the MACHO Photometry Database

    International Nuclear Information System (INIS)

    Alcock, C.; Allsman, R. A.; Alves, D. R.; Axelrod, T. S.; Becker, A. C.; Bennett, D. P.; Cook, K. H.; Drake, A. J.; Freeman, K. C.; Geha, M.

    1999-01-01

    The MACHO Project is a microlensing survey that monitors the brightnesses of ∼60 million stars in the Large Magellanic Cloud (LMC), Small Magellanic Cloud, and Galactic bulge. Our database presently contains about 80 billion photometric measurements, a significant fraction of all astronomical photometry. We describe the calibration of MACHO two-color photometry and transformation to the standard Kron-Cousins V and R system. Calibrated MACHO photometry may be properly compared with all other observations on the Kron-Cousins standard system, enhancing the astrophysical value of these data. For ∼9 million stars in the LMC bar, independent photometric measurements of ∼20,000 stars with V(less-or-similar sign)18 mag in field-overlap regions demonstrate an internal precision σ V =0.021, σ R =0.019, σ V-R =0.028 mag. The accuracy of the zero point in this calibration is estimated to be ±0.035 mag for stars with colors in the range -0.1 mag< V-R<1.2 mag. A comparison of calibrated MACHO photometry with published photometric sequences and new Hubble Space Telescope observations shows agreement. The current calibration zero-point uncertainty for the remainder of the MACHO photometry database is estimated to be ±0.10 mag in V or R and ±0.04 mag in V-R. We describe the first application of calibrated MACHO data: the construction of a color-magnitude diagram used to calculate our experimental sensitivity for detecting microlensing in the LMC. (c) (c) 1999. The Astronomical Society of the Pacific

  2. Calibration of the MACHO Photometry Database

    Energy Technology Data Exchange (ETDEWEB)

    Alcock, C.; Allsman, R. A.; Alves, D. R.; Axelrod, T. S.; Becker, A. C.; Bennett, D. P.; Cook, K. H.; Drake, A. J.; Freeman, K. C.; Geha, M. (and others)

    1999-12-01

    The MACHO Project is a microlensing survey that monitors the brightnesses of {approx}60 million stars in the Large Magellanic Cloud (LMC), Small Magellanic Cloud, and Galactic bulge. Our database presently contains about 80 billion photometric measurements, a significant fraction of all astronomical photometry. We describe the calibration of MACHO two-color photometry and transformation to the standard Kron-Cousins V and R system. Calibrated MACHO photometry may be properly compared with all other observations on the Kron-Cousins standard system, enhancing the astrophysical value of these data. For {approx}9 million stars in the LMC bar, independent photometric measurements of {approx}20,000 stars with V(less-or-similar sign)18 mag in field-overlap regions demonstrate an internal precision {sigma}{sub V} =0.021, {sigma}{sub R} =0.019, {sigma}{sub V-R} =0.028 mag. The accuracy of the zero point in this calibration is estimated to be {+-}0.035 mag for stars with colors in the range -0.1 magcalibrated MACHO photometry with published photometric sequences and new Hubble Space Telescope observations shows agreement. The current calibration zero-point uncertainty for the remainder of the MACHO photometry database is estimated to be {+-}0.10 mag in V or R and {+-}0.04 mag in V-R. We describe the first application of calibrated MACHO data: the construction of a color-magnitude diagram used to calculate our experimental sensitivity for detecting microlensing in the LMC. (c) (c) 1999. The Astronomical Society of the Pacific.

  3. CALIBRATION PROCEDURES ON OBLIQUE CAMERA SETUPS

    Directory of Open Access Journals (Sweden)

    G. Kemper

    2016-06-01

    Full Text Available Beside the creation of virtual animated 3D City models, analysis for homeland security and city planning, the accurately determination of geometric features out of oblique imagery is an important task today. Due to the huge number of single images the reduction of control points force to make use of direct referencing devices. This causes a precise camera-calibration and additional adjustment procedures. This paper aims to show the workflow of the various calibration steps and will present examples of the calibration flight with the final 3D City model. In difference to most other software, the oblique cameras are used not as co-registered sensors in relation to the nadir one, all camera images enter the AT process as single pre-oriented data. This enables a better post calibration in order to detect variations in the single camera calibration and other mechanical effects. The shown sensor (Oblique Imager is based o 5 Phase One cameras were the nadir one has 80 MPIX equipped with a 50 mm lens while the oblique ones capture images with 50 MPix using 80 mm lenses. The cameras are mounted robust inside a housing to protect this against physical and thermal deformations. The sensor head hosts also an IMU which is connected to a POS AV GNSS Receiver. The sensor is stabilized by a gyro-mount which creates floating Antenna –IMU lever arms. They had to be registered together with the Raw GNSS-IMU Data. The camera calibration procedure was performed based on a special calibration flight with 351 shoots of all 5 cameras and registered the GPS/IMU data. This specific mission was designed in two different altitudes with additional cross lines on each flying heights. The five images from each exposure positions have no overlaps but in the block there are many overlaps resulting in up to 200 measurements per points. On each photo there were in average 110 well distributed measured points which is a satisfying number for the camera calibration. In a first

  4. When to Make Mountains out of Molehills: The Pros and Cons of Simple and Complex Model Calibration Procedures

    Science.gov (United States)

    Smith, K. A.; Barker, L. J.; Harrigan, S.; Prudhomme, C.; Hannaford, J.; Tanguy, M.; Parry, S.

    2017-12-01

    Earth and environmental models are relied upon to investigate system responses that cannot otherwise be examined. In simulating physical processes, models have adjustable parameters which may, or may not, have a physical meaning. Determining the values to assign to these model parameters is an enduring challenge for earth and environmental modellers. Selecting different error metrics by which the models results are compared to observations will lead to different sets of calibrated model parameters, and thus different model results. Furthermore, models may exhibit `equifinal' behaviour, where multiple combinations of model parameters lead to equally acceptable model performance against observations. These decisions in model calibration introduce uncertainty that must be considered when model results are used to inform environmental decision-making. This presentation focusses on the uncertainties that derive from the calibration of a four parameter lumped catchment hydrological model (GR4J). The GR models contain an inbuilt automatic calibration algorithm that can satisfactorily calibrate against four error metrics in only a few seconds. However, a single, deterministic model result does not provide information on parameter uncertainty. Furthermore, a modeller interested in extreme events, such as droughts, may wish to calibrate against more low flows specific error metrics. In a comprehensive assessment, the GR4J model has been run with 500,000 Latin Hypercube Sampled parameter sets across 303 catchments in the United Kingdom. These parameter sets have been assessed against six error metrics, including two drought specific metrics. This presentation compares the two approaches, and demonstrates that the inbuilt automatic calibration can outperform the Latin Hypercube experiment approach in single metric assessed performance. However, it is also shown that there are many merits of the more comprehensive assessment, which allows for probabilistic model results, multi

  5. Calibration method for a radwaste assay system

    International Nuclear Information System (INIS)

    Dulama, C.; Dobrin, R.; Toma, Al.; Paunoiu, C.

    2004-01-01

    A waste assay system entirely designed and manufactured in the Institute for Nuclear Research is used in radwaste treatment and conditioning stream to ensure compliance with national repository radiological requirements. Usually, waste assay systems are calibrated by using various experimental arrangements including calibration phantoms. The paper presents a comparative study concerning the efficiency calibration performed by shell source method and a semiempirical, computational method based on a Monte Carlo algorithm. (authors)

  6. Method of Calibrating a Force Balance

    Science.gov (United States)

    Parker, Peter A. (Inventor); Rhew, Ray D. (Inventor); Johnson, Thomas H. (Inventor); Landman, Drew (Inventor)

    2015-01-01

    A calibration system and method utilizes acceleration of a mass to generate a force on the mass. An expected value of the force is calculated based on the magnitude and acceleration of the mass. A fixture is utilized to mount the mass to a force balance, and the force balance is calibrated to provide a reading consistent with the expected force determined for a given acceleration. The acceleration can be varied to provide different expected forces, and the force balance can be calibrated for different applied forces. The acceleration may result from linear acceleration of the mass or rotational movement of the mass.

  7. Calibration and evaluation of neutron moisturemeter

    International Nuclear Information System (INIS)

    Tang Zhangxiong; Hu Jiangchao; Sun Laiyan; Wang Huaihui; Wu Weixue

    1992-02-01

    Factors influencing the calibration curve of neutron moisture meter, such as soil type, texture, volume weight and depth, were studied. When the soil bulk density water content is between 15% to 45%, the calibration curve is approximately a straight line, and the intercept and slope are only influenced by the above factors. The growing plants also influence the calibration curve slightly. The measuring error for top soil (< 20 cm) is larger. The relative error between neutron method and weighing method is about 8%. The neutron method has many advantages such as non-interfering, simple, fast and non-time-delay

  8. Calibration of the ZEUS forward calorimeter

    International Nuclear Information System (INIS)

    Kraemer, M.

    1990-10-01

    The physics at the ep-collider HERA requires high resolution calorimetry calibrated with an accuracy of better than 2%. The ZEUS detector meets these conditions by means of a compensating uranium scintillator sandwich calorimeter with an energy resolution of σ/E = 35%/√E + σ cal , where σ cal is the calibration error. One of the tools to minimize σ cal is the calibration with the signals of the radioactivity of the Uranium plates (UNO). Taking UNO data every 8 hours keeps the calibration stable within ≅ 1%. The muon calibration is done employing an algorithm, that determines the most probable energy loss with a precision of ≅ 1%. The channel-to-channel fluctuations of the ratio μ/UNO for a forward calorimeter (FCAL) prototype show a spread of 5.2% for the electromagnetic calorimeter and ≅ 2.5% for the hadronic sections. Improvements in the construction of the FCAL modules decreased these fluctuations to 2.0% and ≅ 1.8% respectively. The influence of the cracks between the calorimeter modules amounts to ≅ 1.7% on average for the ZEUS geometry, if a 2 mm thick Pb-sheet is introduced between the modules. We conclude that we are able to keep σ cal below 2%. (orig.)

  9. Validating and comparing GNSS antenna calibrations

    Science.gov (United States)

    Kallio, Ulla; Koivula, Hannu; Lahtinen, Sonja; Nikkonen, Ville; Poutanen, Markku

    2018-03-01

    GNSS antennas have no fixed electrical reference point. The variation of the phase centre is modelled and tabulated in antenna calibration tables, which include the offset vector (PCO) and phase centre variation (PCV) for each frequency according to the elevations and azimuths of the incoming signal. Used together, PCV and PCO reduce the phase observations to the antenna reference point. The remaining biases, called the residual offsets, can be revealed by circulating and rotating the antennas on pillars. The residual offsets are estimated as additional parameters when combining the daily GNSS network solutions with full covariance matrix. We present a procedure for validating the antenna calibration tables. The dedicated test field, called Revolver, was constructed at Metsähovi. We used the procedure to validate the calibration tables of 17 antennas. Tables from the IGS and three different calibration institutions were used. The tests show that we were able to separate the residual offsets at the millimetre level. We also investigated the influence of the calibration tables from the different institutions on site coordinates by performing kinematic double-difference baseline processing of the data from one site with different antenna tables. We found small but significant differences between the tables.

  10. A hydrological prediction system based on the SVS land-surface scheme: efficient calibration of GEM-Hydro for streamflow simulation over the Lake Ontario basin

    Directory of Open Access Journals (Sweden)

    É. Gaborit

    2017-09-01

    Full Text Available This work explores the potential of the distributed GEM-Hydro runoff modeling platform, developed at Environment and Climate Change Canada (ECCC over the last decade. More precisely, the aim is to develop a robust implementation methodology to perform reliable streamflow simulations with a distributed model over large and partly ungauged basins, in an efficient manner. The latest version of GEM-Hydro combines the SVS (Soil, Vegetation and Snow land-surface scheme and the WATROUTE routing scheme. SVS has never been evaluated from a hydrological point of view, which is done here for all major rivers flowing into Lake Ontario. Two established hydrological models are confronted to GEM-Hydro, namely MESH and WATFLOOD, which share the same routing scheme (WATROUTE but rely on different land-surface schemes. All models are calibrated using the same meteorological forcings, objective function, calibration algorithm, and basin delineation. GEM-Hydro is shown to be competitive with MESH and WATFLOOD: the NSE  √  (Nash–Sutcliffe criterion computed on the square root of the flows is for example equal to 0.83 for MESH and GEM-Hydro in validation on the Moira River basin, and to 0.68 for WATFLOOD. A computationally efficient strategy is proposed to calibrate SVS: a simple unit hydrograph is used for routing instead of WATROUTE. Global and local calibration strategies are compared in order to estimate runoff for ungauged portions of the Lake Ontario basin. Overall, streamflow predictions obtained using a global calibration strategy, in which a single parameter set is identified for the whole basin of Lake Ontario, show accuracy comparable to the predictions based on local calibration: the average NSE  √  in validation and over seven subbasins is 0.73 and 0.61, respectively for local and global calibrations. Hence, global calibration provides spatially consistent parameter values, robust performance at gauged locations, and reduces the

  11. A hydrological prediction system based on the SVS land-surface scheme: efficient calibration of GEM-Hydro for streamflow simulation over the Lake Ontario basin

    Science.gov (United States)

    Gaborit, Étienne; Fortin, Vincent; Xu, Xiaoyong; Seglenieks, Frank; Tolson, Bryan; Fry, Lauren M.; Hunter, Tim; Anctil, François; Gronewold, Andrew D.

    2017-09-01

    This work explores the potential of the distributed GEM-Hydro runoff modeling platform, developed at Environment and Climate Change Canada (ECCC) over the last decade. More precisely, the aim is to develop a robust implementation methodology to perform reliable streamflow simulations with a distributed model over large and partly ungauged basins, in an efficient manner. The latest version of GEM-Hydro combines the SVS (Soil, Vegetation and Snow) land-surface scheme and the WATROUTE routing scheme. SVS has never been evaluated from a hydrological point of view, which is done here for all major rivers flowing into Lake Ontario. Two established hydrological models are confronted to GEM-Hydro, namely MESH and WATFLOOD, which share the same routing scheme (WATROUTE) but rely on different land-surface schemes. All models are calibrated using the same meteorological forcings, objective function, calibration algorithm, and basin delineation. GEM-Hydro is shown to be competitive with MESH and WATFLOOD: the NSE √ (Nash-Sutcliffe criterion computed on the square root of the flows) is for example equal to 0.83 for MESH and GEM-Hydro in validation on the Moira River basin, and to 0.68 for WATFLOOD. A computationally efficient strategy is proposed to calibrate SVS: a simple unit hydrograph is used for routing instead of WATROUTE. Global and local calibration strategies are compared in order to estimate runoff for ungauged portions of the Lake Ontario basin. Overall, streamflow predictions obtained using a global calibration strategy, in which a single parameter set is identified for the whole basin of Lake Ontario, show accuracy comparable to the predictions based on local calibration: the average NSE √ in validation and over seven subbasins is 0.73 and 0.61, respectively for local and global calibrations. Hence, global calibration provides spatially consistent parameter values, robust performance at gauged locations, and reduces the complexity and computation burden of the

  12. Impact of automatic calibration techniques on HMD life cycle costs and sustainable performance

    Science.gov (United States)

    Speck, Richard P.; Herz, Norman E., Jr.

    2000-06-01

    Automatic test and calibration has become a valuable feature in many consumer products--ranging from antilock braking systems to auto-tune TVs. This paper discusses HMDs (Helmet Mounted Displays) and how similar techniques can reduce life cycle costs and increase sustainable performance if they are integrated into a program early enough. Optical ATE (Automatic Test Equipment) is already zeroing distortion in the HMDs and thereby making binocular displays a practical reality. A suitcase sized, field portable optical ATE unit could re-zero these errors in the Ready Room to cancel the effects of aging, minor damage and component replacement. Planning on this would yield large savings through relaxed component specifications and reduced logistic costs. Yet, the sustained performance would far exceed that attained with fixed calibration strategies. Major tactical benefits can come from reducing display errors, particularly in information fusion modules and virtual `beyond visual range' operations. Some versions of the ATE described are in production and examples of high resolution optical test data will be discussed.

  13. Miniature fission chambers calibration in pulse mode: interlaboratory comparison at the. SCK·CEN BR1 and CEA CALIBAN reactors

    International Nuclear Information System (INIS)

    Lamirand, V.; Geslot, B.; Gregoire, G.; Garnier, D.; Breaud, S.; Mellier, F.; Di-Salvo, J.; Destouches, C.; Blaise, P.; Wagemans, J.; Borms, L.; Malambu, E.; Casoli, P.; Jacquet, X.; Rousseau, G.; Sauvecane, P.

    2013-06-01

    Miniature fission chambers are suited tools for instrumenting experimental reactors, allowing online and in-core neutron measurements of quantities such as fission rates or reactor power. A new set of such detectors was produced by CEA to be used during the next experimental program at the EOLE facility starting in 2013. Some of these detectors will be employed in pulse mode for absolute measurements, thus requiring calibration. The calibration factor is expressed in mass units and thus called 'effective mass'. A calibration campaign was conducted in December 2012 at the SCK.CEN BR1 facility within the framework of the scientific cooperation VEP (VENUS-EOLE-PROTEUS) between SCK.CEN, CEA and PSI. Two actions were conducted in order to improve the calibration method. First a new characterisation of the thermal flux cavity and the MARK3 neutron flux conversion device performed by SCK.CEN allowed using calculated effective cross sections for determining detectors effective masses. Dosimetry irradiations were performed in situ in order to determine the neutron flux level and provide link to the metrological standard. Secondly two fission chambers were also calibrated at the CEA CALIBAN reactor (fast neutron spectrum), using the same method so that the results can be compared with the results obtained at the SCK.CEN. In this paper the calibration method and recent improvements on uncertainty reduction are presented. The results and uncertainties obtained in the two reactors CALIBAN and BR1 are compared and discussed. (authors)

  14. Dosimetry and Calibration Section

    International Nuclear Information System (INIS)

    Otto, T.

    1998-01-01

    The two tasks of the Dosimetry and Calibration Section at CERN are the Individual Dosimetry Service which assures the personal monitoring of about 5000 persons potentially exposed to ionizing radiation at CERN, and the Calibration Laboratory which verifies all the instruments and monitors. This equipment is used by the sections of the RP Group for assuring radiation protection around CERN's accelerators, and by the Environmental Section of TISTE. In addition, nearly 250 electronic and 300 quartz fibre dosimeters, employed in operational dosimetry, are calibrated at least once a year. The Individual Dosimetry Service uses an extended database (INDOS) which contains information about all the individual doses ever received at CERN. For most of 1997 it was operated without the support of a database administrator as the technician who had assured this work retired. The Software Support Section of TIS-TE took over the technical responsibility of the database, but in view of the many other tasks of this Section and the lack of personnel, only a few interventions for solving immediate problems were possible

  15. Phase Calibration of Microphones by Measurement in the Free-field

    Science.gov (United States)

    Shams, Qamar A.; Bartram, Scott M.; Humphreys, William M.; Zuckewar, Allan J.

    2006-01-01

    Over the past several years, significant effort has been expended at NASA Langley developing new Micro-Electro-Mechanical System (MEMS)-based microphone directional array instrumentation for high-frequency aeroacoustic measurements in wind tunnels. This new type of array construction solves two challenges which have limited the widespread use of large channel-count arrays, namely by providing a lower cost-per-channel and a simpler method for mounting microphones in wind tunnels and in field-deployable arrays. The current generation of array instrumentation is capable of extracting accurate noise source location and directivity on a variety of airframe components using sophisticated data reduction algorithms [1-2]. Commercially-available MEMS microphones are condenser-type devices and have some desirable characteristics when compared with conventional condenser-type microphones. The most important advantages of MEMS microphones are their size, price, and power consumption. However, the commercially-available units suffer from certain important shortcomings. Based on experiments with array prototypes, it was found that both the bandwidth and the sound pressure limit of the microphones should be increased significantly to improve the performance and flexibility of the microphone array [3]. It was also desired to modify the packaging to eliminate unwanted Helmholtz resonance s exhibited by the commercial devices. Thus, new requirements were defined as follows: Frequency response: 100 Hz to 100 KHz (+/-3dB) Upper sound pressure limit: Design 1: 130 dB SPL (THD less than 5%) Design 2: 150-160 dB SPL (THD less than 5%) Packaging: 3.73 x 6.13 x 1.3 mm can with laser-etched lid. In collaboration with Novusonic Acoustic Innovation, NASA modified a Knowles SiSonic MEMS design to meet these new requirements. Coupled with the design of the enhanced MEMS microphones was the development of a new calibration method for simultaneously obtaining the sensitivity and phase response of

  16. CAMAC-controlled calibration system for nuclear reactor instruments

    International Nuclear Information System (INIS)

    McDowell, W.P.; Cornella, R.J.

    1977-01-01

    The hardware and the software which have been developed to implement a nuclear instrument calibration system for the Argonne National Laboratory ZPR-VI and ZPR-IX reactor complex are described. The system is implemented using an SEL-840 computer with its associated CAMAC crates and a hardware interface to generate input parameters and measure the required outputs on the instrument under test. Both linear and logarithmic instruments can be calibrated by the system and output parameters can be measured at various automatically selected values of ac line voltage. A complete report on each instrument is printed as a result of the calibration and out-of-tolerance readings are flagged. Operator interface is provided by a CAMAC-controlled Hazeltine terminal. The terminal display leads the operator through the complete calibration procedure. This computer-controlled system is a significant improvement over previously used methods of calibrating nuclear instruments since it reduces reactor downtime and allows rapid detection of long-term changes in instrument calibration

  17. Magnetometer and Gyroscope Calibration Method with Level Rotation

    Directory of Open Access Journals (Sweden)

    Zongkai Wu

    2018-03-01

    Full Text Available Micro electro mechanical system (MEMS gyroscopes and magnetometers are usually integrated into a sensor module or chip and widely used in a variety of applications. In existing integrated gyroscope and magnetometer calibration methods, rotation in all possible orientations is a necessary condition for a good calibration result. However, rotation around two or more axes is difficult to attain, as it is limited by the range of movement of vehicles such as cars, ships, or planes. To solve this problem, this paper proposes an integrated magnetometer and gyroscope calibration method with level rotation. The proposed method presents a redefined magnetometer output model using level attitude. New gyroscope and magnetometer calibration models are then deduced. In addition, a simplified cubature Kalman filter (CKF is established to estimate calibration parameters. This method possesses important value for application in actual systems, as it only needs level rotation for real-time calibration of gyroscopes and magnetometers. Theoretical analysis and test results verify the validity and feasibility of this method.

  18. AVIRIS Inflight Calibration Experiments, Analysis, and Results in 2000

    Science.gov (United States)

    Green, Robert O.; Pavri, Betina

    2001-01-01

    The NASA Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) measures spectra from 370 to 2500 nm with nominally 10-nm sampling and resolution. The spectra are acquired as images with an 11 km width and up to 800 km length from the ER-2 platform or 2.1 km width and 160 km length from the Twin Otter platform. AVIRIS measurements are used for a range of Earth science research and application objectives. The molecular absorption and particle scattering properties expressed in the calibrated AVIRIS measurements are used. For both science research and application objectives, calibration of the AVIRIS spectra is required to produce useful results. Each year prior to the flight season AVIRIS is calibrated in the laboratory. However, the temperature, pressure, vibration, and observational geometry, as well as mechanical, electrical, and operational interfaces of the laboratory are different than the environment on board the airborne platform. To validate the calibration of AVIRIS in the flight environment, an inflight calibration experiment is orchestrated at the beginning of each flight season. In most years additional inflight calibration experiments occur towards the middle and end of the flight season as well. For an inflight calibration experiment, AVIRIS acquires airborne data over a designated calibration target. In concert with the airborne data acquisition, surface and atmospheric properties at the calibration target are measured in situ. These in situ measurements are used to constrain a radiative transfer code and predict the radiance incident at the AVIRIS instrument from the calibration target. This prediction is compared with the AVIRIS-measured radiance to validate the calibration of AVIRIS in the flight environment. Additional properties (such as the AVIRIS inflight measurement precision) are determined as well. This paper presents measurements, analyses, and results from the inflight calibration experiment held on the dry lake bed surface of Rogers Dry

  19. Planck 2015 results: V. LFI calibration

    DEFF Research Database (Denmark)

    Ade, P. A R; Aghanim, N.; Ashdown, M.

    2016-01-01

    We present a description of the pipeline used to calibrate the Planck Low Frequency Instrument (LFI) timelines into thermodynamic temperatures for the Planck 2015 data release, covering four years of uninterrupted operations. As in the 2013 data release, our calibrator is provided by the spin-syn...

  20. Calibration of well-type ionization chambers

    International Nuclear Information System (INIS)

    Alves, C.F.E.; Leite, S.P.; Pires, E.J.; Magalhaes, L.A.G.; David, M.G.; Almeida, C.E. de

    2015-01-01

    This paper presents the methodology developed by the Laboratorio de Ciencias Radiologicas and presently in use for determining of the calibration coefficient for well-type chambers used in the dosimetry of 192 Ir high dose rate sources. Uncertainty analysis involving the calibration procedure are discussed. (author)

  1. Calibration of optical cantilever deflection readers

    International Nuclear Information System (INIS)

    Hu Zhiyu; Seeley, Tim; Kossek, Sebastian; Thundat, Thomas

    2004-01-01

    Because of its ultrahigh sensitivity, the optical lever detection method similar to that used in the atomic force microscope (AFM) has been widely employed as a standard technique for measuring microcantilever deflection. Along with the increasing interest in using the microcantilever as a sensing platform, there is also a requirement for a reliable calibration technique. Many researchers have used the concept of optical lever detection to construct microcantilever deflection readout instruments for chemical, physical, and biological detection. However, without an AFM piezo z scanner, it is very difficult to precisely calibrate these instruments. Here, we present a step-by-step method to conveniently calibrate an instrument using commercially available piezoresistive cantilevers. The experimental results closely match the theoretical calculation. Following this procedure, one can easily calibrate any optical cantilever deflection detection system with high reproducibility, precision, and reliability. A detailed discussion of the optical lever readout system design has been addressed in this article

  2. Improvement of gamma calibration procedures with commercial management software

    International Nuclear Information System (INIS)

    Lucena, Rodrigo F.; Potiens, Maria da Penha A.; Santos, Gelson P.; Vivolo, Vitor

    2007-01-01

    In this work, the gamma calibration procedure of the Instruments Calibration Laboratory (LCI) of the IPEN-CNEN-SP was improved with the use of the commercial management software Autolab TM from Automa Company. That software was adapted for our specific use in the calibration procedures. The evaluation of the uncertainties in gamma calibration protocol was improved by the LCI staff and yet the all worksheets and final calibration report lay-out was developed in commercial software like Excell TM and Word TM from Microsft TM . (author)

  3. ATLAS FCal Diagnostics using the Calibration Pulse

    CERN Document Server

    Rutherfoord, J

    2004-01-01

    The calibration pulser in the ATLAS Forward Calorimeter electronics is used to 1) directly calibrate the warm, active electronics and 2) diagnose the cold, passive electronics chain all the way to the liquid argon electrodes. The study presented here shows that reflections of the calibration pulse coming from discontinuities located at or between the warm preamplifier and the electrode can differentiate and identify all known defects so far observed in this chain.

  4. Dynamic photogrammetric calibration of industrial robots

    Science.gov (United States)

    Maas, Hans-Gerd

    1997-07-01

    Today's developments in industrial robots focus on aims like gain of flexibility, improvement of the interaction between robots and reduction of down-times. A very important method to achieve these goals are off-line programming techniques. In contrast to conventional teach-in-robot programming techniques, where sequences of actions are defined step-by- step via remote control on the real object, off-line programming techniques design complete robot (inter-)action programs in a CAD/CAM environment. This poses high requirements to the geometric accuracy of a robot. While the repeatability of robot poses in the teach-in mode is often better than 0.1 mm, the absolute pose accuracy potential of industrial robots is usually much worse due to tolerances, eccentricities, elasticities, play, wear-out, load, temperature and insufficient knowledge of model parameters for the transformation from poses into robot axis angles. This fact necessitates robot calibration techniques, including the formulation of a robot model describing kinematics and dynamics of the robot, and a measurement technique to provide reference data. Digital photogrammetry as an accurate, economic technique with realtime potential offers itself for this purpose. The paper analyzes the requirements posed to a measurement technique by industrial robot calibration tasks. After an overview on measurement techniques used for robot calibration purposes in the past, a photogrammetric robot calibration system based on off-the- shelf lowcost hardware components will be shown and results of pilot studies will be discussed. Besides aspects of accuracy, reliability and self-calibration in a fully automatic dynamic photogrammetric system, realtime capabilities are discussed. In the pilot studies, standard deviations of 0.05 - 0.25 mm in the three coordinate directions could be achieved over a robot work range of 1.7 X 1.5 X 1.0 m3. The realtime capabilities of the technique allow to go beyond kinematic robot

  5. Traceable calibration of impedance heads and artificial mastoids

    International Nuclear Information System (INIS)

    Scott, D A; Dickinson, L P; Bell, T J

    2015-01-01

    Artificial mastoids are devices which simulate the mechanical characteristics of the human head, and in particular of the bony structure behind the ear. They are an essential tool in the calibration of bone-conduction hearing aids and audiometers. With the emergence of different types of artificial mastoids in the market, and the realisation that the visco-elastic part of these instruments changes over time, the development of a method of traceable calibration of these devices without relying on commercial software has become important for national metrology institutes. This paper describes commercially available calibration methods, and the development of a traceable calibration method including the traceable calibration of the impedance head used to measure the mechanical impedance of the artificial mastoid. (paper)

  6. Optimization of procedure for calibration with radiometer/photometer

    International Nuclear Information System (INIS)

    Detilly, Isabelle

    2009-01-01

    A test procedure for the radiometer/photometer calibrations mark International Light at the Laboratorio de Fotometria y Tecnologia Laser (LAFTA) de la Escuela de Ingenieria Electrica de la Universidad de Costa Rica is established. Two photometric banks are used as experimental set and two calibrations were performed of the International Light. A basic procedure established in the laboratory, is used for calibration from measurements of illuminance and luminous intensity. Some dependent variations of photometric banks used in the calibration process, the programming of the radiometer/photometer and the applied methodology showed the results. The procedure for calibration with radiometer/photometer can be improved by optimizing the programming process of the measurement instrument and possible errors can be minimized by using the recommended procedure. (author) [es

  7. Simple flight time calibration generator in PLL technique

    International Nuclear Information System (INIS)

    Lauch, J.

    1975-01-01

    Calibration and routine check-ups of flight time measuring systems can be carried out with the aid of defined flight time calibration spectra. This paper describes a simple flight time calibration generator capable of generating such calibration spectra in the form of line spectra or of a white spectrum. The flight time of the generator is adjustable in steps from 100 to 3,200 ns. The number of calibration lines can be set to 10 or to 20, resulting in line spacings ranging from 5 to 320 ns. The stop signals are generated by a crystal oscillator, the start signals are generated by a voltage-controlled oscillator locked in a phase control circuit. The start and stop rates can be adjusted in steps. (orig.) [de

  8. Research on orbit prediction for solar-based calibration proper satellite

    Science.gov (United States)

    Chen, Xuan; Qi, Wenwen; Xu, Peng

    2018-03-01

    Utilizing the mathematical model of the orbit mechanics, the orbit prediction is to forecast the space target's orbit information of a certain time based on the orbit of the initial moment. The proper satellite radiometric calibration and calibration orbit prediction process are introduced briefly. On the basis of the research of the calibration space position design method and the radiative transfer model, an orbit prediction method for proper satellite radiometric calibration is proposed to select the appropriate calibration arc for the remote sensor and to predict the orbit information of the proper satellite and the remote sensor. By analyzing the orbit constraint of the proper satellite calibration, the GF-1solar synchronous orbit is chose as the proper satellite orbit in order to simulate the calibration visible durance for different satellites to be calibrated. The results of simulation and analysis provide the basis for the improvement of the radiometric calibration accuracy of the satellite remote sensor, which lays the foundation for the high precision and high frequency radiometric calibration.

  9. Calibration of Spatially Distributed Hydrological Processes and Model Parameters in SWAT Using Remote Sensing Data and an Auto-Calibration Procedure: A Case Study in a Vietnamese River Basin

    Directory of Open Access Journals (Sweden)

    Lan Thanh Ha

    2018-02-01

    Full Text Available In this paper, evapotranspiration (ET and leaf area index (LAI were used to calibrate the SWAT model, whereas remotely sensed precipitation and other climatic parameters were used as forcing data for the 6300 km2 Day Basin, a tributary of the Red River in Vietnam. The efficacy of the Sequential Uncertainty Fitting (SUFI-2 parameter sensitivity and optimization model was tested with area specific remote sensing input parameters for every Hydrological Response Units (HRU, rather than with measurements of river flow representing a large set of HRUs, i.e., a bulk calibration. Simulated monthly ET correlations with remote sensing estimates showed an R2 = 0.71, Nash–Sutcliffe Efficiency NSE = 0.65, and Kling Gupta Efficiency KGE = 0.80 while monthly LAI showed correlations of R2 = 0.59, NSE = 0.57 and KGE = 0.83 over a five-year validation period. Accumulated modelled ET over the 5-year calibration period amounted to 5713 mm compared to 6015 mm of remotely sensed ET, yielding a difference of 302 mm (5.3%. The monthly flow at two flow measurement stations were adequately estimated (R2 = 0.78 and 0.55, NSE = 0.71 and 0.63, KGE = 0.59 and 0.75 for Phu Ly and Ninh Binh, respectively. This outcome demonstrates the capability of SWAT model to obtain spatial and accurate simulation of eco-hydrological processes, also when rivers are ungauged and the water withdrawal system is complex.

  10. Calibration of Geophone Micro-seismic Sensors

    International Nuclear Information System (INIS)

    Bowden, G.

    2005-01-01

    This note describes a simple lab bench technique for calibrating the sensitivity and dynamics of the Mark Products L4C geophones used at SLAC for measuring microseismic ground motion and mechanical vibrations of accelerator support structures. Calibration results for SLAC's three L4C geophones are recorded

  11. Calibration of Ground -based Lidar instrument

    DEFF Research Database (Denmark)

    Villanueva, Héctor; Yordanova, Ginka

    This report presents the result of the lidar calibration performed for the given Ground-based Lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...

  12. Factors affecting the repeatability of gamma camera calibration for quantitative imaging applications using a sealed source

    International Nuclear Information System (INIS)

    Anizan, N; Wahl, R L; Frey, E C; Wang, H; Zhou, X C

    2015-01-01

    Several applications in nuclear medicine require absolute activity quantification of single photon emission computed tomography images. Obtaining a repeatable calibration factor that converts voxel values to activity units is essential for these applications. Because source preparation and measurement of the source activity using a radionuclide activity meter are potential sources of variability, this work investigated instrumentation and acquisition factors affecting repeatability using planar acquisition of sealed sources. The calibration factor was calculated for different acquisition and geometry conditions to evaluate the effect of the source size, lateral position of the source in the camera field-of-view (FOV), source-to-camera distance (SCD), and variability over time using sealed Ba-133 sources. A small region of interest (ROI) based on the source dimensions and collimator resolution was investigated to decrease the background effect. A statistical analysis with a mixed-effects model was used to evaluate quantitatively the effect of each variable on the global calibration factor variability. A variation of 1 cm in the measurement of the SCD from the assumed distance of 17 cm led to a variation of 1–2% in the calibration factor measurement using a small disc source (0.4 cm diameter) and less than 1% with a larger rod source (2.9 cm diameter). The lateral position of the source in the FOV and the variability over time had small impacts on calibration factor variability. The residual error component was well estimated by Poisson noise. Repeatability of better than 1% in a calibration factor measurement using a planar acquisition of a sealed source can be reasonably achieved. The best reproducibility was obtained with the largest source with a count rate much higher than the average background in the ROI, and when the SCD was positioned within 5 mm of the desired position. In this case, calibration source variability was limited by the quantum

  13. Calibration of three-axis magnetometers with differential evolution algorithm

    International Nuclear Information System (INIS)

    Pang, Hongfeng; Zhang, Qi; Wang, Wei; Wang, Junya; Li, Ji; Luo, Shitu; Wan, Chengbiao; Chen, Dixiang; Pan, Mengchun; Luo, Feilu

    2013-01-01

    The accuracy of three-axis magnetometers is influenced by different scale and bias of each axis and nonorthogonality between axes. One limitation of traditional iteration methods is that initial parameters influence the calibration, thus leading to the local optimal or wrong results. In this paper, a new method is proposed to calibrate three-axis magnetometers. To employ this method, a nonmagnetic rotation platform, a proton magnetometer, a DM-050 three-axis magnetometer and the differential evolution (DE) algorithm are used. The performance of this calibration method is analyzed with simulation and experiment. In simulation, the calibration results of DE, unscented Kalman filter (UKF), recursive least squares (RLS) and genetic algorithm (GA) are compared. RMS error using DE is least, which is reduced from 81.233 nT to 1.567 nT. Experimental results show that comparing with UKF, RLS and GA, the DE algorithm has not only the least calibration error but also the best robustness. After calibration, RMS error is reduced from 68.914 nT to 2.919 nT. In addition, the DE algorithm is not sensitive to initial parameters, which is an important advantage compared with traditional iteration algorithms. The proposed algorithm can avoid the troublesome procedure to select suitable initial parameters, thus it can improve the calibration performance of three-axis magnetometers. - Highlights: • The calibration results and robustness of UKF, GA, RLS and DE algorithm are analyzed. • Calibration error of DE is the least in simulation and experiment. • Comparing with traditional calibration algorithms, DE is not sensitive to initial parameters. • It can improve the calibration performance of three-axis magnetometers

  14. Individual dosimetry and calibration

    International Nuclear Information System (INIS)

    Hoefert, M.; Nielsen, M.

    1996-01-01

    In 1995 both the Individual Dosimetry and Calibration Sections worked under the condition of a status quo and concentrated fully on the routine part of their work. Nevertheless, the machine for printing the bar code which will be glued onto the film holder and hence identify the people when entering into high radiation areas was put into operation and most of the holders were equipped with the new identification. As far as the Calibration Section is concerned the project of the new source control system that is realized by the Technical Support Section was somewhat accelerated

  15. Improving calibration accuracy in gel dosimetry

    International Nuclear Information System (INIS)

    Oldham, M.; McJury, M.; Webb, S.; Baustert, I.B.; Leach, M.O.

    1998-01-01

    A new method of calibrating gel dosimeters (applicable to both Fricke and polyacrylamide gels) is presented which has intrinsically higher accuracy than current methods, and requires less gel. Two test-tubes of gel (inner diameter 2.5 cm, length 20 cm) are irradiated separately with a 10x10cm 2 field end-on in a water bath, such that the characteristic depth-dose curve is recorded in the gel. The calibration is then determined by fitting the depth-dose measured in water, against the measured change in relaxivity with depth in the gel. Increased accuracy is achieved in this simple depth-dose geometry by averaging the relaxivity at each depth. A large number of calibration data points, each with relatively high accuracy, are obtained. Calibration data over the full range of dose (1.6-10 Gy) is obtained by irradiating one test-tube to 10 Gy at dose maximum (D max ), and the other to 4.5 Gy at D max . The new calibration method is compared with a 'standard method' where five identical test-tubes of gel were irradiated to different known doses between 2 and 10 Gy. The percentage uncertainties in the slope and intercept of the calibration fit are found to be lower with the new method by a factor of about 4 and 10 respectively, when compared with the standard method and with published values. The gel was found to respond linearly within the error bars up to doses of 7 Gy, with a slope of 0.233±0.001 s -1 Gy -1 and an intercept of 1.106±0.005 Gy. For higher doses, nonlinear behaviour was observed. (author)

  16. Portable and fixed monitoring units for tank calibrations and monitoring of process liquids

    International Nuclear Information System (INIS)

    Landat, D.A.; Hunt, B.A.

    1999-01-01

    The development work stems from safeguards support activities carried out at the JRC Ispra, Italy to the inspectorate agencies. A range of measurement equipment covering the needs of the inspector have been designed, developed and tested in both the laboratory and in nuclear facilities. The instruments comprise four units: (1) a portable pressure measurement device, (2) a volume long term monitoring device, (3) an unattended volume measurement system and (4) a level measurement unit. Utilization of the equipment has proven to give independent measurement checks and confirmation of operator's instrumentation and declarations, ensuring continuity of knowledge. (J.P.N.)

  17. Construction of calibration curve for accountancy tank

    International Nuclear Information System (INIS)

    Kato, Takayuki; Goto, Yoshiki; Nidaira, Kazuo

    2009-01-01

    Tanks are equipped in a reprocessing plant for accounting solution of nuclear material. The careful measurement of volume in tanks is very important to implement rigorous accounting of nuclear material. The calibration curve relating the volume and level of solution needs to be constructed, where the level is determined by differential pressure of dip tubes. Several calibration curves are usually employed, but it's not explicitly decided how many segment are used, where to select segment, or what should be the degree of polynomial curve. These parameters, i.e., segment and degree of polynomial curve are mutually interrelated to give the better performance of calibration curve. Here we present the construction technique of giving optimum calibration curves and their characteristics. (author)

  18. Calibration of the neutron scintillation counter threshold

    International Nuclear Information System (INIS)

    Noga, V.I.; Ranyuk, Yu.N.; Telegin, Yu.N.

    1978-01-01

    A method for calibrating the threshold of a neutron counter in the form of a 10x10x40 cm plastic scintillator is described. The method is based on the evaluation of the Compton boundary of γ-spectrum from the discrimination curve of counter loading. The results of calibration using 60 Co and 24 Na γ-sources are given. In order to eValuate the Compton edge rapidly, linear extrapolation of the linear part of the discrimination curve towards its intersection with the X axis is recommended. Special measurements have shown that the calibration results do not practically depend on the distance between the cathode of a photomultiplier and the place where collimated γ-radiation of the calibration source reaches the scintillator

  19. Evaluation of methods to calibrate radiation survey meters

    International Nuclear Information System (INIS)

    Robinson, R.C.; Arbeau, N.D.

    1987-04-01

    Calibration requirements for radiation survey meters used in industrial radiography have been reviewed. Information obtained from a literature search, discussions with CSLD inspectors and firms performing calibrations has been considered. Based on this review a set of minimum calibration requirements was generated which, when met, will determine that the survey meter is suited for measurements described in the current AEC Regulations that apply to industrial radiography equipment. These requirements are presented in this report and may be used as guidelines for evaluating calibration methods proposed or in use in industry. 39 refs

  20. Calibration of ionization chambers used in LDR brachytherapy

    International Nuclear Information System (INIS)

    Alvarez, Oscar T.B.; Caldas, Linda V.E.

    2005-01-01

    In this work was developed a calibration procedure of well-type ionization chambers used for measurements of I-125, seed type. It was used as a standard an ionization chamber Capintec CRC-15BT, with calibration certificate of the University of Wisconsin. It were calibrated two well-type ionization chambers of Capintec CRC-15R model. A source of I-125 was used in clinical use (18.5 to 7.4 MBq). The results showed that with the application of calibration factors was possible to decrease read deviate from 16% to just 1.0%