WorldWideScience

Sample records for temperature measurement methods

  1. Noise thermometry - a new temperature measuring method

    International Nuclear Information System (INIS)

    Brixy, H.; Hecker, R.; Rittinghaus, K.F.

    1975-01-01

    The thermal Johnson-Niquist noise is the basis of noise thermometry. This temperature measuring method is, e.g., of interest insofar as the noise thermometer gives absolute values as a primary thermometer and is in principle extensively independent of environmental influences and material properties. The resistance values of the measuring probe are about 10 Ohm to a few kOhm. The demands of electronics are high, the self-noise of the measuring apparatus must be as small as possible; a comparative measuring method is advantageous. 1 to 2,500 K are given as a possible temperature range. An accuracy of 0.1% could be achieved in laboratory measurements. Temperature measurements to be used in operation in a few nuclear reactors are mentioned. (HP/LH) [de

  2. Two methods to measure granular gas temperature

    Science.gov (United States)

    Chastaing, J.-Y.; Géminard, J.-C.; Naert, A.

    2017-07-01

    Grains are vibrated so as to achieve a granular gas, here regarded as an archetype of a dissipative non equilibrium steady state (NESS). We report on two distinct and concordant experimental measures of the system effective temperature. To do so, a blade fastened to the shaft of a small DC-motor, immersed in the grains, behaves as a driven 1D Brownian rotator, which is used as both actuator and sensor simultaneously. On the one hand, the Gallavotti-Cohen fluctuation theorem, which involves a measure of the asymmetry of the energy exchanges between the rotator and the NESS reservoir, provides a first effective temperature. On the other hand, the fluctuation-dissipation theorem, which involves the relation between the spontaneous fluctuations and the response to a weak perturbation, defines a second, independent, effective temperature. Both methods, even though they are based on drastically different ideas, give nicely concordant results.

  3. New methode of measurement of temperature flow

    Directory of Open Access Journals (Sweden)

    Slávka Grexová

    2008-03-01

    Full Text Available The subject of this article is a measurement of thermal flow under laboratory conditions. We define the thermal flow as an amount of heat transmitted through a surface of rock over a certain period of time. According to the Atlas of Geothermal Energy, thermal flow ranges from 40 to 120 mW/m2. It is not possible to measure it directly on the rock surface. The conventional ways of measurement is a “separation bar” thermic conduction measurement system or measurement of the temperature of the rock in two different places at selected underground depth intervals. These measurements and analyses are not sufficient to make a final conclusion. It is necessary to repeat the measurements under real conditions.

  4. Liquid temperature measuring method and device therefor

    International Nuclear Information System (INIS)

    Maruyama, Fumi; Karasawa, Hirokazu.

    1995-01-01

    In the present invention, temperature of liquid metal in coolants in an FBR type reactor can accurately be measured at rapid response time. Namely, ultrasonic waves are emitted from an ultrasonic wave sensor disposed in the air to a guide wave tube. Ultrasonic waves are reflected at reflection plates disposed at front and back or upper and lower portions of a small hole disposed to the wave guide tube. The reflected waves are received by the sensor described above. The difference of the reaching time of the reflected waves from the reflecting plates disposed at the front and the back or the upper and lower portions is measured. The speed of sounds in this case is determined based on the size of the small hole and the distance of the upper and the lower reflection plates. The speed of sounds is determined by the formula below: V(m/s) = 2500 - 0.52 T, where T: temperature. The temperature of the liquid can easily be calculated based on the formula. Accordingly, since the speed of the ultrasonic waves from their emission to the reception is msec order, and the processing of the signals are simple, the temperature can be measured at a response time of several msecs. In addition, since the ultrasonic wave sensor is disposed at the outside of the reactor, no special countermeasure for environmental circumstances is necessary, to improve maintenance ability. (I.S.)

  5. Wet method for measuring starch gelatinization temperature using electrical conductivity.

    Science.gov (United States)

    Morales-Sanchez, E; Figueroa, J D C; Gaytan-Martínez, M

    2009-09-01

    The objective of the present study was to develop a method for obtaining the gelatinization temperature of starches by using electrical conductivity. Native starches from corn, rice, potato, and wheat were prepared with different proportions of water and heated from room temperature to 90 degrees C, in a device especially designed for monitoring the electrical conductivity as a function of temperature. The results showed a linear trend of the electrical conductivity with the temperature until it reaches the onset gelatinization temperature. After that point, the electrical conductivity presented an increment or decrement depending on the water content in the sample and it was related to starch swelling and gelatinization phenomena. At the end gelatinization temperature, the conductivity becomes stable and linear, indicating that there are no more changes of phase. The starch gelatinization parameter, which was evaluated in the 4 types of starches using the electrical conductivity, was compared with those obtained by using differential scanning calorimeter (DSC). The onset temperature at which the electrical conductivity increased or decreased was found to be similar to that obtained by DSC. Also, the final temperature at which the electrical conductivity returned to linearity matched the end gelatinization temperature of the DSC. Further, a wet method for measuring the onset, peak, and end gelatinization temperatures as a function of temperature using the electrical conductivity curves is presented for a starch-water suspension.

  6. Fiber-optical method of pyrometric measurement of melts temperature

    Science.gov (United States)

    Zakharenko, V. A.; Veprikova, Ya R.

    2018-01-01

    There is a scientific problem of non-contact measurement of the temperature of metal melts now. The problem is related to the need to achieve the specified measurement errors in conditions of uncertainty of the blackness coefficients of the radiating surfaces. The aim of this work is to substantiate the new method of measurement in which the influence of the blackness coefficient is eliminated. The task consisted in calculating the design and material of special crucible placed in the molten metal, which is an emitter in the form of blackbody (BB). The methods are based on the classical concepts of thermal radiation and calculations based on the Planck function. To solve the problem, the geometry of the crucible was calculated on the basis of the Goofy method which forms the emitter of a blackbody at the immersed in the melt. The paper describes the pyrometric device based on fiber optic pyrometer for temperature measurement of melts, which implements the proposed method of measurement using a special crucible. The emitter is formed by the melt in this crucible, the temperature within which is measured by means of fiber optic pyrometer. Based on the results of experimental studies, the radiation coefficient ε‧ > 0.999, which confirms the theoretical and computational justification is given in the article

  7. Pyrometric method for measuring emittances at high temperatures

    Science.gov (United States)

    Ballestrín, J.; Rodríguez, J.; Carra, M. E.; Cañadas, I.; Roldan, M. I.; Barbero, J.; Marzo, A.

    2016-05-01

    In this work an alternative method for emittance determination based on pyrometric measurements is presented. The measurement procedure has been applied to AISI 310S steel samples in the Plataforma Solar de Almería vertical axis solar furnace SF5. The experimental results show that emittance increases with increasing temperature and decreases with increasing wavelength. This behaviour is in agreement with experimental results obtained by other authors. Analysis of tests has revealed a good repeatability (1%) and accuracy (< 2%) of this measurement procedure.

  8. A comparison of four methods of normal newborn temperature measurement.

    Science.gov (United States)

    Sganga, A; Wallace, R; Kiehl, E; Irving, T; Witter, L

    2000-01-01

    The purpose of this study was to: (a) compare newborn temperature measurements obtained by digital disposable, electronic, and tympanic thermometers with glass mercury thermometers, and (b) compare financial implications of each method. In this correlational study, 12 perinatal and neonatal nurses obtained temperature measurements of 184 newborns between 1 and 168 hours of age. The stratified convenience sample was selected using medical records numbers. Temperature instruments included glass thermometer, tympanic thermometer, electronic thermometer, and a digital thermometer. Data were analyzed by Pearson r coefficients, mean, standard deviation, and range using an SPSS statistical package. The glass thermometer, electronic thermometer, and digital thermometer temperature assessments were highly correlated (0.748-1.0). The tympanic thermometer had a low correlation coefficient (0.35). Use of the glass thermometer had the highest accompanying cost. Tympanic thermometers were the most cost effective. In healthy newborns, the use of electronic and digital thermometers can be encouraged if there is concern about using glass thermometers. These results cannot be extrapolated to sick infants. While tympanic thermometers had the lowest associated cost, their lack of correlation with the gold standard glass thermometers for accurate temperature assessment makes them a poor choice for healthy newborns.

  9. Methods of Temperature and Emission Measure Determination of Coronal Loops

    Science.gov (United States)

    Cirtain, J. W.; Schmelz, J. T.; Martens, P. C. H.

    2002-05-01

    Recent observational results from both SOHO-EIT and TRACE indicate that coronal loops are isothermal along their length (axially). These results are obtained from a narrowband filter ratio method that assumes that the plasma is isothermal along the line of sight (radially). However, these temperatures vary greatly from those derived from differential emission measure (DEM) curves produced from spectral lines recorded by SOHO-CDS. The DEM results indicate that the loops are neither axially nor radially isothermal. This discrepancy was investigated by Schmelz et al. (2001). They chose pairs of iron lines from the same CDS data set to mimic the EIT and TRACE loop results. Ratios of different lines gave different temperatures, indicating that the plasma was not radially isothermal. In addition the results indicated that the loop was axially isothermal, even though the DEM analysis of the same data showed this result to be false. Here we have analyzed the EIT data for the CDS loop published by Schmelz et al. (2001). We took the ratios of the 171-to-195 and 195-to-284 filter data, and made temperature maps of the loop. The results indicate that the loop is axially isothermal, but different temperatures were found for each pair of filters. Both ratio techniques force the resultant temperature to lie within the range where the response functions (for filters) or the emissivity functions (for lines) overlap; isothermal loops are therefore a byproduct of the analysis. This conclusion strengthens support for the idea that temperature and emission measure results from filter ratio methods may be misleading or even drastically wrong. This research was funded in part by the NASA/TRACE MODA grant for Montana State University. Solar physics research at the University of Memphis is supported by NASA grant NAG5-9783.

  10. [Comparison of different methods of temperature measurment in children].

    Science.gov (United States)

    Pavlović, Momcilo; Radlović, Nedeljko; Leković, Zoran; Berenji, Karolina

    2008-01-01

    The consequences of failing to notice fever in children can be serious. On the other hand, false positive reading can result in unnecesery investigation or diagnostic approach. The aim of this study was to compare different ways of body temperature measurement. This prospective study was carried out on Pediatric Department of General Hospital in Subotica during 10 months (March-December 2006). In 263 children aged 1 month to 18 years of age, the body temperature was obtained from 4 measurement sites: tactile assesment, forehead and ear by electronic thermometer, rectal temperature in small children (up to 2 years of age) or axillar temperature in older children by mercury thermometer. Tympanic thermometry was considered as a standard for fever detection. The sensitivity of rectal temperature to detect fever is 46.67%, while specificity is 92.19%. The sensitivity of fever detection by electronic thermometry on the forehead is lower according to rectal thermometry - 36.08%, while specificity is 95.18%. The lowest values ofsensitivity are recorded in axillar thermometry (35.82%), specificity is 90.20%. The correlation coefficient is higher between tympanic and rectal temperature measurement (r=0.5076, pchildren and tympanic thermometry in children over 2 years of age.

  11. Temperature measurement

    Science.gov (United States)

    ... this page: //medlineplus.gov/ency/article/003400.htm Temperature measurement To use the sharing features on this page, please enable JavaScript. The measurement of body temperature can help detect illness. It can also monitor ...

  12. A new computer method for temperature measurement based on an optimal control problem

    NARCIS (Netherlands)

    Damean, N.; Houkes, Z.; Regtien, Paulus P.L.

    1996-01-01

    A new computer method to measure extreme temperatures is presented. The method reduces the measurement of the unknown temperature to the solving of an optimal control problem, using a numerical computer. Based on this method, a new device for temperature measurement is built. It consists of a

  13. Temperature measurement of burning aluminum powder based on the double line method of atomic emission spectra

    Science.gov (United States)

    Tang, Huijuan; Hao, Xiaojian; Hu, Xiaotao

    2018-01-01

    In the case of conventional contact temperature measurement, there is a delay phenomenon and high temperature resistant materials limitation. By using the faster response speed and theoretically no upper limit of the non-contact temperature method, the measurement system based on the principle of double line atomic emission spectroscopy temperature measurement is put forward, the structure and theory of temperature measuring device are introduced. According to the atomic spectrum database (ASD), Aluminum(Al) I 690.6 nm and Al I 708.5 nm are selected as the two lines in the temperature measurement. The intensity ratio of the two emission lines was measured by a spectrometer to obtain the temperature of Al burning in pure oxygen, and the result compared to the temperature measured by the thermocouple. It turns out that the temperature correlation between the two methods is good, and it proves the feasibility of the method.

  14. Evaluation of methods for gravity wave extraction from middle atmospheric lidar temperature measurements

    OpenAIRE

    Ehard, Benedikt; Kaifler, Bernd; Kaifler, Natalie; Rapp, Markus

    2015-01-01

    This study evaluates commonly used methods of extracting gravity wave induced temperature perturbations from lidar measurements. The spectral response of these methods is characterized with the help of a synthetic dataset with known temperature perturbations added to a realistic background temperature profile. The simulations are carried out with the background temperature being either constant or varying in time to evaluate the sensitivit...

  15. Development of a method for estimating oesophageal temperature by multi-locational temperature measurement inside the external auditory canal

    Science.gov (United States)

    Nakada, Hirofumi; Horie, Seichi; Kawanami, Shoko; Inoue, Jinro; Iijima, Yoshinori; Sato, Kiyoharu; Abe, Takeshi

    2017-09-01

    We aimed to develop a practical method to estimate oesophageal temperature by measuring multi-locational auditory canal temperatures. This method can be applied to prevent heatstroke by simultaneously and continuously monitoring the core temperatures of people working under hot environments. We asked 11 healthy male volunteers to exercise, generating 80 W for 45 min in a climatic chamber set at 24, 32 and 40 °C, at 50% relative humidity. We also exposed the participants to radiation at 32 °C. We continuously measured temperatures at the oesophagus, rectum and three different locations along the external auditory canal. We developed equations for estimating oesophageal temperatures from auditory canal temperatures and compared their fitness and errors. The rectal temperature increased or decreased faster than oesophageal temperature at the start or end of exercise in all conditions. Estimated temperature showed good similarity with oesophageal temperature, and the square of the correlation coefficient of the best fitting model reached 0.904. We observed intermediate values between rectal and oesophageal temperatures during the rest phase. Even under the condition with radiation, estimated oesophageal temperature demonstrated concordant movement with oesophageal temperature at around 0.1 °C overestimation. Our method measured temperatures at three different locations along the external auditory canal. We confirmed that the approach can credibly estimate the oesophageal temperature from 24 to 40 °C for people performing exercise in the same place in a windless environment.

  16. An Improved Tumour Temperature Measurement and Control Method for Superficial Tumour Ultrasound Hyperthermia Therapeutic System

    Energy Technology Data Exchange (ETDEWEB)

    Shen, G F; Chen, Y Z; Ren, G X [Biomedical Instrument Institute, Shanghai Jiao Tong University, Shanghai 200030 (China)

    2006-10-15

    In tumour hyperthermia therapy, the research on measurement and control of tumour temperature is very important. Based on the hardware platform of superficial tumour ultrasound hyperthermia therapeutic system, an improved tumour temperature measurement and control method is presented in this paper. The experiment process, data and results are discussed in detail. The improved method will greatly reduce the pain and dread of the patients during the therapy period on the tumour temperature measurement and control by using the pinhead sensor.

  17. An experimental method for making spectral emittance and surface temperature measurements of opaque surfaces

    International Nuclear Information System (INIS)

    Moore, Travis J.; Jones, Matthew R.; Tree, Dale R.; Daniel Maynes, R.; Baxter, Larry L.

    2011-01-01

    An experimental procedure has been developed to make spectral emittance and temperature measurements. The spectral emittance of an object is calculated using measurements of the spectral emissive power and of the surface temperature of the object obtained using a Fourier transform infrared (FTIR) spectrometer. A calibration procedure is described in detail which accounts for the temperature dependence of the detector. The methods used to extract the spectral emissive power and surface temperature from measured infrared spectra were validated using a blackbody radiator at known temperatures. The average error in the measured spectral emittance was 2.1% and the average difference between the temperature inferred from the recorded spectra and the temperature indicated on the blackbody radiator was 1.2%. The method was used to measure the spectral emittance of oxidized copper at various temperatures.

  18. Measurement of surface temperature and emissivity by a multitemperature method for Fourier-transform infrared spectrometers

    DEFF Research Database (Denmark)

    Clausen, Sønnik; Morgenstjerne, Axel; Rathmann, Ole

    1996-01-01

    measurement at a known sample temperature, for example, at ambient temperature. The temperature of the sample surface can be measured rather easily at ambient temperature. The spectrum at ambient temperature is used to eliminate background effects from spectra as measured at other surface temperatures....... The temperatures of the sample are found in a single calculation from the measured spectra independently of the response function of the instrument and the emissivity of the sample. The spectral emissivity of a sample can be measured if the instrument is calibrated against a blackbody source. Temperatures...... of blackbody sources are estimated with an uncertainty of 0.2-2 K. The method is demonstrated for measuring the spectral emissivity of a brass specimen and an oxidized nickel specimen. (C) 1996 Optical Society of America...

  19. Method for independent strain and temperature measurement in polymeric tensile test specimen using embedded FBG sensors

    DEFF Research Database (Denmark)

    Pereira, Gilmar Ferreira; McGugan, Malcolm; Mikkelsen, Lars Pilgaard

    2016-01-01

    to calculate independently the strain and temperature are presented in the article, together with a measurement resolution study. This multi-parameter measurement method was applied to an epoxy tensile specimen, tested in a unidirectional tensile test machine with a temperature controlled cabinet. A full......A novel method to obtain independent strain and temperature measurements using embedded Fibre Bragg Grating (FBG) in polymeric tensile test specimens is presented in this paper. The FBG strain and temperature cross-sensitivity was decoupled using two single mode FBG sensors, which were embedded...... of temperature, from 40 C to -10 C. The consistency of the expected theoretical results with the calibration procedure and the experimental validation shows that this proposed method is applicable to measure accurate strain and temperature in polymers during static or fatigue tensile testing. Two different...

  20. Coiling Temperature Control Using Temperature Measurement Method for the Hot Rolled Strip in the Water Cooling Banks

    Science.gov (United States)

    Nakagawa, Shigemasa; Tachibana, Hisayoshi; Honda, Tatsuro; Uematsu, Chihiro

    In the hot strip mill, the quality of the strip greatly depends on the cooling process between the last stand in the finishing mill and the coilers. Therefore, it is important to carefully control the coiling temperature to regulate the mechanical properties of the strip. To realize high accuracy of coiling temperature, a new coiling temperature control using temperature measurement method for the hot rolled strip in the water cooling banks has been developed. The features of the new coiling temperature control are as follows: (i) New feedforward control adjusts ON/OFF swiching of cooling headers according to the strip temperature measured in the water cooling banks. (ii) New feedforward control is achieved by dynamic control function. This coiling temperature control has been in operation successfully since 2008 at Kashima Steel Works and improved the accuracy of coiling temperature of high strength steel considerably.

  1. Evaluation of methods for gravity wave extraction from middle-atmospheric lidar temperature measurements

    Directory of Open Access Journals (Sweden)

    B. Ehard

    2015-11-01

    Full Text Available This study evaluates commonly used methods of extracting gravity-wave-induced temperature perturbations from lidar measurements. The spectral response of these methods is characterized with the help of a synthetic data set with known temperature perturbations added to a realistic background temperature profile. The simulations are carried out with the background temperature being either constant or varying in time to evaluate the sensitivity to temperature perturbations not caused by gravity waves. The different methods are applied to lidar measurements over New Zealand, and the performance of the algorithms is evaluated. We find that the Butterworth filter performs best if gravity waves over a wide range of periods are to be extracted from lidar temperature measurements. The running mean method gives good results if only gravity waves with short periods are to be analyzed.

  2. A new optical method for measuring surface temperature at large incident probe angles

    Science.gov (United States)

    Lee, A. S.; Norris, P. M.

    1997-02-01

    A novel thermoreflectance technique has been developed for noncontact temperature measurements using laser light incident at large angles on solid materials and devices. The method involves measuring the differential reflectance from a polarization modulated laser beam. The polarization differential reflectance technique is demonstrated on single-crystal Si wafers and on a polycrystalline carbon thin film over a temperature range of 20-60 °C. The method is shown to be an extremely sensitive temperature probe for near grazing angle measurements, which could be useful for monitoring the surface temperature of closely stacked silicon wafers used in batch processing in the microelectronics industry.

  3. A method enabling simultaneous pressure and temperature measurement using a single piezoresistive MEMS pressure sensor

    International Nuclear Information System (INIS)

    Frantlović, Miloš; Stanković, Srđan; Jokić, Ivana; Lazić, Žarko; Smiljanić, Milče; Obradov, Marko; Vukelić, Branko; Jakšić, Zoran

    2016-01-01

    In this paper we present a high-performance, simple and low-cost method for simultaneous measurement of pressure and temperature using a single piezoresistive MEMS pressure sensor. The proposed measurement method utilizes the parasitic temperature sensitivity of the sensing element for both pressure measurement correction and temperature measurement. A parametric mathematical model of the sensor was established and its parameters were calculated using the obtained characterization data. Based on the model, a real-time sensor correction for both pressure and temperature measurements was implemented in a target measurement system. The proposed method was verified experimentally on a group of typical industrial-grade piezoresistive sensors. The obtained results indicate that the method enables the pressure measurement performance to exceed that of typical digital industrial pressure transmitters, achieving at the same time the temperature measurement performance comparable to industrial-grade platinum resistance temperature sensors. The presented work is directly applicable in industrial instrumentation, where it can add temperature measurement capability to the existing pressure measurement instruments, requiring little or no additional hardware, and without adverse effects on pressure measurement performance. (paper)

  4. Research on position calibration method in infrared scanning temperature measurement system of rotary kiln

    Science.gov (United States)

    Dai, Shao-sheng; You, Chang-hui; Guo, Zhong-yuan; Cheng, Ya-jun; Yu, Liang-bing

    2016-11-01

    Aiming at the large error in the equal-interval locating method, a precise position calibration method is proposed. The proposed method improves the location measurement accuracy by introducing some feature temperature points to divide the rotary kiln into several segments, then the equal-interval locating method was applied to each segment, ultimately, a position calibration data more closing to the actual situation was got. The feature temperature points can be selected from the temperature points of kiln tyres or the highest temperature point and so on. Taking the practical application into consideration, the best result is obtained, when four feature temperature points was introduced to divide the rotary kiln into five segments. The experiment result shows that compared with the equal-interval method, the accuracy of the proposed method has raised about 5.6 times when four feature temperature points is used.

  5. Deformation measurements of materials at low temperatures using laser speckle photography method

    International Nuclear Information System (INIS)

    Sumio Nakahara; Yukihide Maeda; Kazunori Matsumura; Shigeyoshi Hisada; Takeyoshi Fujita; Kiyoshi Sugihara

    1992-01-01

    The authors observed deformations of several materials during cooling down process from room temperature to liquid nitrogen temperature using the laser speckle photography method. The in-plane displacements were measured by the image plane speckle photography and the out-of-plane displacement gradients by the defocused speckle photography. The results of measurements of in-plane displacement are compared with those of FEM analysis. The applicability of laser speckle photography method to cryogenic engineering are also discussed

  6. Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients.

    Science.gov (United States)

    Asadian, Simin; Khatony, Alireza; Moradi, Gholamreza; Abdi, Alireza; Rezaei, Mansour

    2016-01-01

    An accurate determination of body temperature in critically ill patients is a fundamental requirement for initiating the proper process of diagnosis, and also therapeutic actions; therefore, the aim of the study was to assess the accuracy and precision of four noninvasive peripheral methods of temperature measurement compared to the central nasopharyngeal measurement. In this observational prospective study, 237 patients were recruited from the intensive care unit of Imam Ali Hospital of Kermanshah. The patients' body temperatures were measured by four peripheral methods; oral, axillary, tympanic, and forehead along with a standard central nasopharyngeal measurement. After data collection, the results were analyzed by paired t-test, kappa coefficient, receiver operating characteristic curve, and using Statistical Package for the Social Sciences, version 19, software. There was a significant meaningful correlation between all the peripheral methods when compared with the central measurement (Ptemperatures of right and left tympanic membranes and the standard central nasopharyngeal measurement (88%). Paired t-test demonstrated an acceptable precision with forehead (P=0.132), left (P=0.18) and right (P=0.318) tympanic membranes, oral (P=1.00), and axillary (P=1.00) methods. Sensitivity and specificity of both the left and right tympanic membranes were more than for other methods. The tympanic and forehead methods had the highest and lowest accuracy for measuring body temperature, respectively. It is recommended to use the tympanic method (right and left) for assessing a patient's body temperature in the intensive care units because of high accuracy and acceptable precision.

  7. Method for local temperature measurement in a nanoreactor for in situ high-resolution electron microscopy.

    Science.gov (United States)

    Vendelbo, S B; Kooyman, P J; Creemer, J F; Morana, B; Mele, L; Dona, P; Nelissen, B J; Helveg, S

    2013-10-01

    In situ high-resolution transmission electron microscopy (TEM) of solids under reactive gas conditions can be facilitated by microelectromechanical system devices called nanoreactors. These nanoreactors are windowed cells containing nanoliter volumes of gas at ambient pressures and elevated temperatures. However, due to the high spatial confinement of the reaction environment, traditional methods for measuring process parameters, such as the local temperature, are difficult to apply. To address this issue, we devise an electron energy loss spectroscopy (EELS) method that probes the local temperature of the reaction volume under inspection by the electron beam. The local gas density, as measured using quantitative EELS, is combined with the inherent relation between gas density and temperature, as described by the ideal gas law, to obtain the local temperature. Using this method we determined the temperature gradient in a nanoreactor in situ, while the average, global temperature was monitored by a traditional measurement of the electrical resistivity of the heater. The local gas temperatures had a maximum of 56 °C deviation from the global heater values under the applied conditions. The local temperatures, obtained with the proposed method, are in good agreement with predictions from an analytical model. Copyright © 2013 Elsevier B.V. All rights reserved.

  8. Kinetic methods for measuring the temperature of clusters and nanoparticles in molecular beams

    International Nuclear Information System (INIS)

    Makarov, Grigorii N

    2011-01-01

    The temperature (internal energy) of clusters and nanoparticles is an important physical parameter which affects many of their properties and the character of processes they are involved in. At the same time, determining the temperature of free clusters and nanoparticles in molecular beams is a rather complicated problem because the temperature of small particles depends on their size. In this paper, recently developed kinetic methods for measuring the temperature of clusters and nanoparticles in molecular beams are reviewed. The definition of temperature in the present context is given, and how the temperature affects the properties of and the processes involving the particles is discussed. The temperature behavior of clusters and nanoparticles near a phase transition point is analyzed. Early methods for measuring the temperature of large clusters are briefly described. It is shown that, compared to other methods, new kinetic methods are more universal and applicable for determining the temperature of clusters and nanoparticles of practically any size and composition. The future development and applications of these methods are outlined. (reviews of topical problems)

  9. Accuracy and precision of four common peripheral temperature measurement methods in intensive care patients

    Directory of Open Access Journals (Sweden)

    Asadian S

    2016-09-01

    Full Text Available Simin Asadian,1 Alireza Khatony,1 Gholamreza Moradi,2 Alireza Abdi,1 Mansour Rezaei,3 1Nursing and Midwifery School, Kermanshah University of Medical Sciences, 2Department of Anesthesiology, 3Biostatistics & Epidemiology Department, Kermanshah University of Medical Sciences, Kermanshah, Iran Introduction: An accurate determination of body temperature in critically ill patients is a fundamental requirement for initiating the proper process of diagnosis, and also therapeutic actions; therefore, the aim of the study was to assess the accuracy and precision of four noninvasive peripheral methods of temperature measurement compared to the central nasopharyngeal measurement. Methods: In this observational prospective study, 237 patients were recruited from the intensive care unit of Imam Ali Hospital of Kermanshah. The patients’ body temperatures were measured by four peripheral methods; oral, axillary, tympanic, and forehead along with a standard central nasopharyngeal measurement. After data collection, the results were analyzed by paired t-test, kappa coefficient, receiver operating characteristic curve, and using Statistical Package for the Social Sciences, version 19, software. Results: There was a significant meaningful correlation between all the peripheral methods when compared with the central measurement (P<0.001. Kappa coefficients showed good agreement between the temperatures of right and left tympanic membranes and the standard central nasopharyngeal measurement (88%. Paired t-test demonstrated an acceptable precision with forehead (P=0.132, left (P=0.18 and right (P=0.318 tympanic membranes, oral (P=1.00, and axillary (P=1.00 methods. Sensitivity and specificity of both the left and right tympanic membranes were more than for other methods. Conclusion: The tympanic and forehead methods had the highest and lowest accuracy for measuring body temperature, respectively. It is recommended to use the tympanic method (right and left for

  10. Prediction of human core body temperature using non-invasive measurement methods

    Science.gov (United States)

    Niedermann, Reto; Wyss, Eva; Annaheim, Simon; Psikuta, Agnes; Davey, Sarah; Rossi, René Michel

    2014-01-01

    The measurement of core body temperature is an efficient method for monitoring heat stress amongst workers in hot conditions. However, invasive measurement of core body temperature (e.g. rectal, intestinal, oesophageal temperature) is impractical for such applications. Therefore, the aim of this study was to define relevant non-invasive measures to predict core body temperature under various conditions. We conducted two human subject studies with different experimental protocols, different environmental temperatures (10 °C, 30 °C) and different subjects. In both studies the same non-invasive measurement methods (skin temperature, skin heat flux, heart rate) were applied. A principle component analysis was conducted to extract independent factors, which were then used in a linear regression model. We identified six parameters (three skin temperatures, two skin heat fluxes and heart rate), which were included for the calculation of two factors. The predictive value of these factors for core body temperature was evaluated by a multiple regression analysis. The calculated root mean square deviation (rmsd) was in the range from 0.28 °C to 0.34 °C for all environmental conditions. These errors are similar to previous models using non-invasive measures to predict core body temperature. The results from this study illustrate that multiple physiological parameters (e.g. skin temperature and skin heat fluxes) are needed to predict core body temperature. In addition, the physiological measurements chosen in this study and the algorithm defined in this work are potentially applicable as real-time core body temperature monitoring to assess health risk in broad range of working conditions.

  11. Prediction of human core body temperature using non-invasive measurement methods.

    Science.gov (United States)

    Niedermann, Reto; Wyss, Eva; Annaheim, Simon; Psikuta, Agnes; Davey, Sarah; Rossi, René Michel

    2014-01-01

    The measurement of core body temperature is an efficient method for monitoring heat stress amongst workers in hot conditions. However, invasive measurement of core body temperature (e.g. rectal, intestinal, oesophageal temperature) is impractical for such applications. Therefore, the aim of this study was to define relevant non-invasive measures to predict core body temperature under various conditions. We conducted two human subject studies with different experimental protocols, different environmental temperatures (10 °C, 30 °C) and different subjects. In both studies the same non-invasive measurement methods (skin temperature, skin heat flux, heart rate) were applied. A principle component analysis was conducted to extract independent factors, which were then used in a linear regression model. We identified six parameters (three skin temperatures, two skin heat fluxes and heart rate), which were included for the calculation of two factors. The predictive value of these factors for core body temperature was evaluated by a multiple regression analysis. The calculated root mean square deviation (rmsd) was in the range from 0.28 °C to 0.34 °C for all environmental conditions. These errors are similar to previous models using non-invasive measures to predict core body temperature. The results from this study illustrate that multiple physiological parameters (e.g. skin temperature and skin heat fluxes) are needed to predict core body temperature. In addition, the physiological measurements chosen in this study and the algorithm defined in this work are potentially applicable as real-time core body temperature monitoring to assess health risk in broad range of working conditions.

  12. Measurement of the body surface temperature by the method of laser photothermal radiometry

    International Nuclear Information System (INIS)

    Skvortsov, L A; Kirillov, V M

    2003-01-01

    The specific features of contactless measurements of the body surface temperature by the method of repetitively pulsed laser photothermal radiometry are considered and the requirements to the parameters of the laser and measurement scheme are formulated. The sensitivity of the method is estimated. The advantages of laser photothermal radiometry over the conventional passive radiometric method are discussed. (laser applications and other topics in quantum electronics)

  13. Thermal Expansion and Magnetostriction Measurements at Cryogenic Temperature Using the Strain Gauge Method.

    Science.gov (United States)

    Wang, Wei; Liu, Huiming; Huang, Rongjin; Zhao, Yuqiang; Huang, Chuangjun; Guo, Shibin; Shan, Yi; Li, Laifeng

    2018-01-01

    Thermal expansion and magnetostriction, the strain responses of a material to temperature and a magnetic field, especially properties at low temperature, are extremely useful to study electronic and phononic properties, phase transitions, quantum criticality, and other interesting phenomena in cryogenic engineering and materials science. However, traditional dilatometers cannot provide magnetic field and ultra-low temperature (thermal expansion and magnetostriction at cryogenic temperature using the strain gauge method based on a Physical Properties Measurements System (PPMS). The interfacing software and automation were developed using LabVIEW. The sample temperature range can be tuned continuously between 1.8 and 400 K. With this PPMS-aided measuring system, we can observe temperature and magnetic field dependence of the linear thermal expansion of different solid materials easily and accurately.

  14. Thermal Expansion and Magnetostriction Measurements at Cryogenic Temperature Using the Strain Gauge Method

    Directory of Open Access Journals (Sweden)

    Wei Wang

    2018-03-01

    Full Text Available Thermal expansion and magnetostriction, the strain responses of a material to temperature and a magnetic field, especially properties at low temperature, are extremely useful to study electronic and phononic properties, phase transitions, quantum criticality, and other interesting phenomena in cryogenic engineering and materials science. However, traditional dilatometers cannot provide magnetic field and ultra-low temperature (<77 K environment easily. This paper describes the design and test results of thermal expansion and magnetostriction at cryogenic temperature using the strain gauge method based on a Physical Properties Measurements System (PPMS. The interfacing software and automation were developed using LabVIEW. The sample temperature range can be tuned continuously between 1.8 and 400 K. With this PPMS-aided measuring system, we can observe temperature and magnetic field dependence of the linear thermal expansion of different solid materials easily and accurately.

  15. An investigation of temperature measurement methods in nuclear power plant reactor pressure vessel annealing

    International Nuclear Information System (INIS)

    Acton, R.U.; Gill, W.; Sais, D.J.; Schulze, D.H.; Nakos, J.T.

    1996-05-01

    The objective of this project was to provide an assessment of several methods by which the temperature of a commercial nuclear power plant reactor pressure vessel (RPV) could be measured during an annealing process. This project was a coordinated effort between DOE's Office of Nuclear Energy, Science and Technology; DOE's Light Water Reactor Technology Center at Sandia National Laboratories; and the Electric Power Research Institute's Non- Destructive Evaluation Center. Ball- thermocouple probes similar to those described in NUREG/CR-5760, spring-loaded, metal- sheathed thermocouple probes, and 1778 air- suspended thermocouples were investigated in experiments that heated a section of an RPV wall to simulate a thermal annealing treatment. A parametric study of ball material, emissivity, thermal conductivity, and thermocouple function locations was conducted. Also investigated was a sheathed thermocouple failure mode known as shunting (electrical breakdown of insulation separating the thermocouple wires). Large errors were found between the temperature as measured by the probes and the true RPV wall temperature during heat-up and cool-down. At the annealing soak temperature, in this case 454 degrees C [850'F], all sensors measured the same temperature within about ±5% (23.6 degrees C [42.5 degrees F]). Because of these errors, actual RPV wall heating and cooling rates differed from those prescribed (by up to 29%). Shunting does not appear to be a problem under these conditions. The large temperature measurement errors led to the development of a thermal model that predicts the RPV wall temperature from the temperature of a ball- probe. Comparisons between the model and the experimental data for ball-probes indicate that the model could be a useful tool in predicting the actual RPV temperature based on the indicated ball- probe temperature. The model does not predict the temperature as well for the spring-loaded and air suspended probes

  16. Research on calibration method of downhole optical fiber temperature measurement and its application in SAGD well

    Science.gov (United States)

    Lu, Zhiwei; Han, Li; Hu, Chengjun; Pan, Yong; Duan, Shengnan; Wang, Ningbo; Li, Shijian; Nuer, Maimaiti

    2017-10-01

    With the development of oil and gas fields, the accuracy and quantity requirements of real-time dynamic monitoring data needed for well dynamic analysis and regulation are increasing. Permanent, distributed downhole optical fiber temperature and pressure monitoring and other online real-time continuous data monitoring has become an important data acquisition and transmission technology in digital oil field and intelligent oil field construction. Considering the requirement of dynamic analysis of steam chamber developing state in SAGD horizontal wells in F oil reservoir in Xinjiang oilfield, it is necessary to carry out real-time and continuous temperature monitoring in horizontal section. Based on the study of the principle of optical fiber temperature measurement, the factors that cause the deviation of optical fiber temperature sensing are analyzed, and the method of fiber temperature calibration is proposed to solve the problem of temperature deviation. Field application in three wells showed that it could attain accurate measurement of downhole temperature by temperature correction. The real-time and continuous downhole distributed fiber temperature sensing technology has higher application value in the reservoir management of SAGD horizontal wells. It also has a reference for similar dynamic monitoring in reservoir production.

  17. Method and system to measure temperature of gases using coherent anti-stokes doppler spectroscopy

    Science.gov (United States)

    Rhodes, Mark

    2013-12-17

    A method of measuring a temperature of a noble gas in a chamber includes providing the noble gas in the chamber. The noble gas is characterized by a pressure and a temperature. The method also includes directing a first laser beam into the chamber and directing a second laser beam into the chamber. The first laser beam is characterized by a first frequency and the second laser beam is characterized by a second frequency. The method further includes converting at least a portion of the first laser beam and the second laser beam into a coherent anti-Stokes beam, measuring a Doppler broadening of the coherent anti-Stokes beam, and computing the temperature using the Doppler broadening.

  18. A fully automated temperature-dependent resistance measurement setup using van der Pauw method

    Science.gov (United States)

    Pandey, Shivendra Kumar; Manivannan, Anbarasu

    2018-03-01

    The van der Pauw (VDP) method is widely used to identify the resistance of planar homogeneous samples with four contacts placed on its periphery. We have developed a fully automated thin film resistance measurement setup using the VDP method with the capability of precisely measuring a wide range of thin film resistances from few mΩ up to 10 GΩ under controlled temperatures from room-temperature up to 600 °C. The setup utilizes a robust, custom-designed switching network board (SNB) for measuring current-voltage characteristics automatically at four different source-measure configurations based on the VDP method. Moreover, SNB is connected with low noise shielded coaxial cables that reduce the effect of leakage current as well as the capacitance in the circuit thereby enhancing the accuracy of measurement. In order to enable precise and accurate resistance measurement of the sample, wide range of sourcing currents/voltages are pre-determined with the capability of auto-tuning for ˜12 orders of variation in the resistances. Furthermore, the setup has been calibrated with standard samples and also employed to investigate temperature dependent resistance (few Ω-10 GΩ) measurements for various chalcogenide based phase change thin films (Ge2Sb2Te5, Ag5In5Sb60Te30, and In3SbTe2). This setup would be highly helpful for measurement of temperature-dependent resistance of wide range of materials, i.e., metals, semiconductors, and insulators illuminating information about structural change upon temperature as reflected by change in resistances, which are useful for numerous applications.

  19. Reconstruction method for inversion problems in an acoustic tomography based temperature distribution measurement

    International Nuclear Information System (INIS)

    Liu, Sha; Liu, Shi; Tong, Guowei

    2017-01-01

    In industrial areas, temperature distribution information provides a powerful data support for improving system efficiency, reducing pollutant emission, ensuring safety operation, etc. As a noninvasive measurement technology, acoustic tomography (AT) has been widely used to measure temperature distribution where the efficiency of the reconstruction algorithm is crucial for the reliability of the measurement results. Different from traditional reconstruction techniques, in this paper a two-phase reconstruction method is proposed to ameliorate the reconstruction accuracy (RA). In the first phase, the measurement domain is discretized by a coarse square grid to reduce the number of unknown variables to mitigate the ill-posed nature of the AT inverse problem. By taking into consideration the inaccuracy of the measured time-of-flight data, a new cost function is constructed to improve the robustness of the estimation, and a grey wolf optimizer is used to solve the proposed cost function to obtain the temperature distribution on the coarse grid. In the second phase, the Adaboost.RT based BP neural network algorithm is developed for predicting the temperature distribution on the refined grid in accordance with the temperature distribution data estimated in the first phase. Numerical simulations and experiment measurement results validate the superiority of the proposed reconstruction algorithm in improving the robustness and RA. (paper)

  20. Reconstruction method for inversion problems in an acoustic tomography based temperature distribution measurement

    Science.gov (United States)

    Liu, Sha; Liu, Shi; Tong, Guowei

    2017-11-01

    In industrial areas, temperature distribution information provides a powerful data support for improving system efficiency, reducing pollutant emission, ensuring safety operation, etc. As a noninvasive measurement technology, acoustic tomography (AT) has been widely used to measure temperature distribution where the efficiency of the reconstruction algorithm is crucial for the reliability of the measurement results. Different from traditional reconstruction techniques, in this paper a two-phase reconstruction method is proposed to ameliorate the reconstruction accuracy (RA). In the first phase, the measurement domain is discretized by a coarse square grid to reduce the number of unknown variables to mitigate the ill-posed nature of the AT inverse problem. By taking into consideration the inaccuracy of the measured time-of-flight data, a new cost function is constructed to improve the robustness of the estimation, and a grey wolf optimizer is used to solve the proposed cost function to obtain the temperature distribution on the coarse grid. In the second phase, the Adaboost.RT based BP neural network algorithm is developed for predicting the temperature distribution on the refined grid in accordance with the temperature distribution data estimated in the first phase. Numerical simulations and experiment measurement results validate the superiority of the proposed reconstruction algorithm in improving the robustness and RA.

  1. A review of the methods to measure the ion temperature in a tokamak plasma

    International Nuclear Information System (INIS)

    Zurro Hernandez, B.; Perez-Navarro Gomez, A.

    1976-01-01

    The most important methods to measure the ion temperatu--re in a Tokamak plasma are reviewed, e.g. energy analysis of the fast neutrals which leave out the plasma, Doppler broadening of the emision spectral lines and fusion neutron analysis. It is discussed their bounds so as the advantages and drawbacks of each one. Other methods of some interest in the future are outlined. (author) [es

  2. Calorimetric method for current sharing temperature measurements in ITER conductor samples in SULTAN

    International Nuclear Information System (INIS)

    Bagnasco, M.

    2009-01-01

    Several Toroidal Field Conductor short samples with slight layout variations have been assembled and tested in the SULTAN facility at CRPP. The measurement campaigns started in 2007 and are still ongoing. The performance of every conductor is expressed in terms of current sharing temperature (T cs ), i.e. the temperature at which a defined electric field, 10 μV/m, is detected in the cable due to the incipient superconducting-to-normal state transition. The T cs at specific operating conditions is the key design parameter for the ITER conductors and is the main object of the qualification tests. Typically, the average electric field is measured with voltage tap pairs attached on the jacket along the conductor. The inability however to explain observed premature voltage developments opened the discussion about possible alternative measuring methods. The He flow calorimetric method is based on the measurement of the resistive power generation in the conductor. It relies on the detection of very small temperature increases along the conductor in steady state operation. The accuracy and the reliability of the calorimetric method in SULTAN are critically discussed, with particular emphasis on the instrumentation requirements and test procedures. The application of the calorimetric method to the recent SULTAN test campaigns is described with its merits and limits. For future tests of ITER conductors in SULTAN, the calorimetric method for T cs test is proposed as a routine procedure.

  3. Development of the CARS method for measurement of pressure and temperature gradients in centrifuges

    International Nuclear Information System (INIS)

    Zeltmann, A.H.; Valentini, J.J.

    1983-12-01

    These experiments evaluated the feasibility of applying the CARS technique to the measurement of UF 6 concentrations and pressure gradients in a gas centrifuge. The resultant CARS signals were properly related to system parameters as suggested by theory. The results have been used to guide design of an apparatus for making CARS measurements in a UF 6 gas centrifuge. Ease of measurement is expected for pressures as low as 0.1 torr. Temperature gradients can be measured by this technique with changes in the data acquisition method. 16 references, 8 figures, 2 tables

  4. Modified T-history method for measuring thermophysical properties of phase change materials at high temperature

    Science.gov (United States)

    Omaraa, Ehsan; Saman, Wasim; Bruno, Frank; Liu, Ming

    2017-06-01

    Latent heat storage using phase change materials (PCMs) can be used to store large amounts of energy in a narrow temperature difference during phase transition. The thermophysical properties of PCMs such as latent heat, specific heat and melting and solidification temperature need to be defined at high precision for the design and estimating the cost of latent heat storage systems. The existing laboratory standard methods, such as differential thermal analysis (DTA) and differential scanning calorimetry (DSC), use a small sample size (1-10 mg) to measure thermophysical properties, which makes these methods suitable for homogeneous elements. In addition, this small amount of sample has different thermophysical properties when compared with the bulk sample and may have limitations for evaluating the properties of mixtures. To avoid the drawbacks in existing methods, the temperature - history (T-history) method can be used with bulk quantities of PCM salt mixtures to characterize PCMs. This paper presents a modified T-history setup, which was designed and built at the University of South Australia to measure the melting point, heat of fusion, specific heat, degree of supercooling and phase separation of salt mixtures for a temperature range between 200 °C and 400 °C. Sodium Nitrate (NaNO3) was used to verify the accuracy of the new setup.

  5. A steady-state high-temperature method for measuring thermal conductivity of refractory materials

    Science.gov (United States)

    Manzolaro, M.; Corradetti, S.; Andrighetto, A.; Ferrari, L.

    2013-05-01

    A new methodology and an instrumental setup for the thermal conductivity estimation of isotropic bulk graphite and different carbides at high temperatures are presented. The method proposed in this work is based on the direct measurement of temperature and emissivity on the top surface of a sample disc of known dimensions. Temperatures measured under steady-state thermal equilibrium are then used to estimate the thermal conductivity of the sample by making use of the inverse parameter estimation technique. Thermal conductivity values obtained in this way are then compared to the material data sheets and values found in literature. The reported work has been developed within the Research and Development framework of the SPES (Selective Production of Exotic Species) project at INFN-LNL (Istituto Nazionale di Fisica Nucleare - Laboratori Nazionali di Legnaro).

  6. Radiation temperature measurement method for semitransparent materials using one-channel infrared pyrometer.

    Science.gov (United States)

    Fu, Tairan; Liu, Jiangfan; Zong, Anzhou

    2014-10-10

    Semitransparent zinc sulfide (ZnS) crystal materials are widely used as the infrared-transmitting windows for optical instruments operating in long wavelengths. This paper describes a temperature measurement method for high-temperature ZnS materials using the one-channel optical pyrometer based on a theoretical model of radiation transfer in semitransparent plates. Numerical analyses of the radiation properties of ZnS plate are used to optimize the spectral band for the optical pyrometry. The optimized measurement spectral band is based on a trade-off between the measurement radiation intensity and the signal-to-noise ratio (SNR) for the ZnS material. The effective waveband emittance of one-dimensional (1D) ZnS plates is analyzed for various experimental conditions (temperatures, thicknesses, and direction angles) for the one-channel infrared pyrometer with the optimized measurement spectral response. The analysis can be used to improve radiation temperature measurements of semitransparent ZnS materials in applications.

  7. METHODS AND TECHNIQUE FOR THERMOPOWER AND ELECTRICAL CONDUCTIVITY MEASUREMENTS OF THERMOELECTRIC MATERIALS AT HIGH TEMPERATURES

    Directory of Open Access Journals (Sweden)

    A. T. Burkov

    2015-02-01

    Full Text Available The principles and methods of thermopower and electrical conductivity measurements at high temperatures (100 – 1000 K are reviewed. These two properties define the so-called power factor of thermoelect ric materials. Moreover, in combination with thermal conductivity they give the efficiency of thermoelectric conversion. In spite of a principal obtained by different groups and hinders a realistic estimate of the potential thermoelectric efficiency of new materials. The lack of a commonly accepted reference material for thermopower exaggerates the problem. Therefore, it is very important to have a clear understanding of the capabilities and limitations of the measuring methods and set-ups. The review article deals with the definitions of the thermoelectric parameters and principles of their experimental determination. The metrological characteristics of the state-of-the-art experimental set-ups for high temperature measurements are analyzed. simplicity of the measurement methods of these properties, their practical realization is rather complicated, especially at high temperatures. This leads to large uncertainties in determination of the properties, complicates comparison of the results

  8. Development of a new method for the noninvasive measurement of deep body temperature without a heater.

    Science.gov (United States)

    Kitamura, Kei-Ichiro; Zhu, Xin; Chen, Wenxi; Nemoto, Tetsu

    2010-01-01

    The conventional zero-heat-flow thermometer, which measures the deep body temperature from the skin surface, is widely used at present. However, this thermometer requires considerable electricity to power the electric heater that compensates for heat loss from the probe; thus, AC power is indispensable for its use. Therefore, this conventional thermometer is inconvenient for unconstrained monitoring. We have developed a new dual-heat-flux method that can measure the deep body temperature from the skin surface without a heater. Our method is convenient for unconstrained and long-term measurement because the instrument is driven by a battery and its design promotes energy conservation. Its probe consists of dual-heat-flow channels with different thermal resistances, and each heat-flow-channel has a pair of IC sensors attached on its top and bottom. The average deep body temperature measurements taken using both the dual-heat-flux and then the zero-heat-flow thermometers from the foreheads of 17 healthy subjects were 37.08 degrees C and 37.02 degrees C, respectively. In addition, the correlation coefficient between the values obtained by the 2 methods was 0.970 (ppower supply. (c) 2009 IPEM. Published by Elsevier Ltd. All rights reserved.

  9. Electron temperature distribution measurement in Z pinch by the laser radiation absorption method

    International Nuclear Information System (INIS)

    Sarkisov, G.S.; Ehtlisher, B.

    1995-01-01

    The paper proposes a new method for measuring the temperature of internal region of Z-pinch; the method is based on laser probing of plasma and simultaneous measurements of absorption profiles and interferential run-on of probing radiation phase inside Z-pinch. It is shown that opaque pinch region which occurs laser probing experiments is related to the absorption of radiation. It is shown that laser radiation which has passed through Z-pinch features sufficiently high degree of coherence which enables interferometry of absorption region. The values of electron density and electron temperature were recovered in the experiments with Z-pinches (produced after an explosion of 20 μm aluminium wire by 50 ns pulse of current of about 250 kA for necking the pinch of ∼ 1.4 x 10 20 cm -3 and ∼ 530 eV, respectively. 11 refs., 4 figs

  10. Verifying the distributed temperature sensing Bowen ratio method for measuring evaporation

    Science.gov (United States)

    Schilperoort, Bart; Coenders-Gerrits, Miriam; Luxemburg, Willem; Cisneros Vaca, César; Ucer, Murat

    2016-04-01

    Evaporation is an important process in the hydrological cycle, therefore measuring evaporation accurately is essential for water resource management, hydrological management and climate change models. Current techniques to measure evaporation, like eddy covariance systems, scintillometers, or lysimeters, have their limitations and therefore cannot always be used to estimate evaporation correctly. Also the conventional Bowen ratio surface energy balance method has as drawback that two sensors are used, which results in large measuring errors. In Euser et al. (2014) a new method was introduced, the DTS-based Bowen ratio (BR-DTS), that overcomes this drawback. It uses a distributed temperature sensing technique (DTS) whereby a fibre optic cable is placed vertically, going up and down along a measurement tower. One stretch of the cable is dry, the other wrapped with cloth and kept wet, akin to a psychrometer. Using this, the wet and dry bulb temperatures are determined every 12.5 cm over the height, from which the Bowen ratio can be determined. As radiation and wind have an effect on the cooling and heating of the cable's sheath as well, the DTS cables do not necessarily always measure dry and wet bulb temperature of the air accurately. In this study the accuracy in representing the dry and wet bulb temperatures of the cable are verified, and evaporation observations of the BR-DTS method are compared to Eddy Covariance (EC) measurements. Two ways to correct for errors due to wind and solar radiation warming up the DTS cables are presented: one for the dry cable and one for the wet cable. The measurements were carried out in a pine forest near Garderen (The Netherlands), along a 46-meter tall scaffold tower (15 meters above the canopy). Both the wet (Twet) and dry (Tdry) temperature of the DTS cable were compared to temperature and humidity (from which Twet is derived) observations from sensors placed along the height of the tower. Underneath the canopy, where there was

  11. The Elastic Constants Measurement of Metal Alloy by Using Ultrasonic Nondestructive Method at Different Temperature

    Directory of Open Access Journals (Sweden)

    Eryi Hu

    2016-01-01

    Full Text Available The ultrasonic nondestructive method is introduced into the elastic constants measurement of metal material. The extraction principle of Poisson’s ratio, elastic modulus, and shear modulus is deduced from the ultrasonic propagating equations with two kinds of vibration model of the elastic medium named ultrasonic longitudinal wave and transverse wave, respectively. The ultrasonic propagating velocity is measured by using the digital correlation technique between the ultrasonic original signal and the echo signal from the bottom surface, and then the elastic constants of the metal material are calculated. The feasibility of the correlation algorithm is verified by a simulation procedure. Finally, in order to obtain the stability of the elastic properties of different metal materials in a variable engineering application environment, the elastic constants of two kinds of metal materials in different temperature environment are measured by the proposed ultrasonic method.

  12. Operating temperature measuring method for SnO2 gas-sensing materials using infra-red sensor

    Science.gov (United States)

    Liang, Yu; Sun, Yongquan; Wu, Tong; Zhang, Jing

    2017-10-01

    Operating temperature was crucial for SnO2 gas sensor considering the serious impacts on sensors' selectivity and reliability. While, it was difficult to measure this operating temperature because the size of the sensitive body was small, as well as its heat capacity. In this paper, the temperature signal was acquired by the non-contact infrared temperature sensor and processed by the signal conditioning circuit and single-chip, and then the measured temperature were displayed by the single-chip. The method of subsection calibration was adopted to improve the accuracy of temperature measurement. Finally, the uncertainty of system measurement was estimated.

  13. Double modulation pyrometry: A radiometric method to measure surface temperatures of directly irradiated samples

    Science.gov (United States)

    Potamias, Dimitrios; Alxneit, Ivo; Wokaun, Alexander

    2017-09-01

    The design, implementation, calibration, and assessment of double modulation pyrometry to measure surface temperatures of radiatively heated samples in our 1 kW imaging furnace is presented. The method requires that the intensity of the external radiation can be modulated. This was achieved by a rotating blade mounted parallel to the optical axis of the imaging furnace. Double modulation pyrometry independently measures the external radiation reflected by the sample as well as the sum of thermal and reflected radiation and extracts the thermal emission as the difference of these signals. Thus a two-step calibration is required: First, the relative gains of the measured signals are equalized and then a temperature calibration is performed. For the latter, we transfer the calibration from a calibrated solar blind pyrometer that operates at a different wavelength. We demonstrate that the worst case systematic error associated with this procedure is about 300 K but becomes negligible if a reasonable estimate of the sample's emissivity is used. An analysis of the influence of the uncertainties in the calibration coefficients reveals that one (out of the five) coefficient contributes almost 50% to the final temperature error. On a low emission sample like platinum, the lower detection limit is around 1700 K and the accuracy typically about 20 K. Note that these moderate specifications are specific for the use of double modulation pyrometry at the imaging furnace. It is mainly caused by the difficulty to achieve and maintain good overlap of the hot zone with a diameter of about 3 mm Full Width at Half Height and the measurement spot both of which are of similar size.

  14. A Simple Method For The Time-Resolved Measurement Of Electron Temperature In A Z-pinch

    International Nuclear Information System (INIS)

    Robledo-Martinez, Arturo

    2006-01-01

    A new method for measuring the electron temperature in a Z-pinch plasma with time resolution is presented here. PIN diodes fitted with different filters were employed to detect the X-radiation in separate energy bands. By comparison of the relative intensities of the X-ray pulses, a time-resolved measurement of the electron temperature can be obtained. The first step of the implementation of the method consists in convoluting a continuum spectrum with the filter-detector response to estimate individual signal strength. The second step is to calculate the ratio of the amplitudes of 2 differently-filtered diodes as a function of temperature to obtain a temperature calibration. When the computed calibration is then compared with the experimental signal ratio the temperature can be obtained. The temperatures measured using the present method were found to have peak values from 3 to 7 keV. The method only applies when the plasma is emitting continuum radiation

  15. Strain and temperature measurement

    International Nuclear Information System (INIS)

    Stewart, P.A.E.; Fowler, P.H.

    1988-01-01

    A method of non-invasively measuring strain and temperature of an object, substantially simultaneously, using neutrons of selected energy levels is described. A pulsed neutron source is made to emit thermal and epithermal neutrons in a collimated beam directed at the object. Temperature is monitored by observing the thermal Doppler broadening of resonances in the neutron transmission characteristic for the epithermal neutrons and strain is measured from observations made of changes to the thermal neutron diffraction pattern. The object may be a gas turbine blade or a thrust bearing. (author)

  16. Equipment for temperature measurements

    OpenAIRE

    HORELICA, Josef

    2008-01-01

    This work deals with theoretical description, practical connection of temperature measuring instruments and how they function while measuring the temperature. This document explaines basic concepts of temperature, temperature scales and temperature measuring. Further there are introduced standard types of sensors used in these measuring intruments. This document includes a picture presentation in PowerPoint, where an aggregate table of sensors and an application of these sensors in measuring ...

  17. A spectroscopic temperature measurement of converging detonations by the emission spectra-matching method

    International Nuclear Information System (INIS)

    Sugimura, Tadayoshi; Fujiwara, Toshitaka.

    1980-01-01

    The spectroscopic measurement of the temperature of converging detonation by the emission spectrum-matching method was proposed and performed. The combination of gas dynamics parameters was adjusted until the agreement between the calculated profile of artificial spectra and the light emission spectra at the convergence center was obtained. The mixed gas of oxygen and acetylene was used for the experiment. When the mixing ratio of oxygen and acetylene was one to one, and the initial gas pressure was 60 Torr, the behavior of detonation was same as that of cylindrically converging shock waves propagating in the gas of specific heat ratio of 1.28 in the Guderley's theory. The increase of light emission intensity with the increase of molecular density, and the effect of temperature were observed. The effect of pressure broadening was also studied. It was found that the spectrum-matching method is available even for the weak emission or strong broadening. The artificial light emission spectra which agreed with the observed spectra were obtained. (Kato, T.)

  18. Total hemispherical emittance measured at high temperatures by the calorimetric method

    International Nuclear Information System (INIS)

    DiFilippo, F.; Mirtich, M.J.; Banks, B.A.; Stidham, C.; Kussmaul, M.

    1994-01-01

    A calorimetric vacuum emissometer (CVE) capable of measuring total hemispherical emittance of surfaces at elevated temperatures was designed, built, and tested. Several materials with a wide range of emittances were measured in the CVE between 773 to 923 K. These results were compared to values calculated from spectral emittance curves measured in a room temperature Hohlraum reflectometer and in an open-air elevated temperature emissometer. The results differed by as much as 0.2 for some materials but were in closer agreement for the more highly-emitting, diffuse-reflecting samples. The differences were attributed to temperature, atmospheric, and directional effects, and errors in the Hohlraum and emissometer measurements (± 5 percent). The probable error of the CVE measurements was typically less than 1 percent

  19. Effective neutron temperature measurements in well moderated reactor by the reactivity coefficient method

    International Nuclear Information System (INIS)

    Raisic, N.; Klinc, T.

    1968-11-01

    The ratio of the reactivity changes of a nuclear reactor produced by successive introduction of two different neutron absorbers in the reactor core, has been measured and information on effective neutron temperature at a particular point obtained. Boron was used as a l/v absorber and cadmium as an absorber sensiti ve to neutron temperature. Effective neutron temperature distribution has been deduced by moving absorbers across the reactor core and observing the corresponding reactivity changes. (author)

  20. A method for measuring pH at high temperatures is presented

    International Nuclear Information System (INIS)

    Chaudon, Luc.

    1979-01-01

    Two hydrogen electrodes are used and set up in a PTFE cell comprising two chambers connected through a saturated potassium chloride solution bridge. This cell is put in an autoclave containing hydrogen. The potential difference of the following cell is measured: H 2 , Pt, R solution - KCl saturated solution at 25 0 C - X solution, Pt, H 2 - The pH of the reference solution R is known up to 300 0 C and the X solution must have its pH to be determined. The precision of the measures at 300 0 C is estimated about +-0,1 pH unit. The dissociation constant of water is calculated from pH variations of alcaline solutions up to 300 0 C. The method has helped to measure the pH at 300 0 C of some boric acid solutions, with or without lithium hydroxide additions, in the following concentration range: B: 250 to 1500 ppm and Li: 0 to 3 ppm. Some concentrations are in fact those chosen for the primary circuits of pressurized water reactors. The pH of ammoniacal solutions is measured too and helped to determine the variations of the dissociation constant of ammonia with temperature [fr

  1. A new method to measure Bowen ratios using high-resolution vertical dry and wet bulb temperature profiles

    NARCIS (Netherlands)

    Euser, T.; Luxemburg, W.M.J.; Everson, C.S.; Mengistu, M.G.; Clulow, A.D.; Bastiaanssen, W.G.M.

    2014-01-01

    The Bowen ratio surface energy balance method is a relatively simple method to determine the latent heat flux and the actual land surface evaporation. The Bowen ratio method is based on the measurement of air temperature and vapour pressure gradients. If these measurements are performed at only two

  2. Arctic (and Antarctic) Observing Experiment - an Assessment of Methods to Measure Temperature over Polar Environments

    Science.gov (United States)

    Rigor, I. G.; Clemente-Colon, P.; Nghiem, S. V.; Hall, D. K.; Woods, J. E.; Henderson, G. R.; Zook, J.; Marshall, C.; Gallage, C.

    2014-12-01

    The Arctic environment has been undergoing profound changes; the most visible is the dramatic decrease in Arctic sea ice extent (SIE). These changes pose a challenge to our ability to measure surface temperature across the Polar Regions. Traditionally, the International Arctic Buoy Programme (IABP) and International Programme for Antarctic Buoys (IPAB) have measured surface air temperature (SAT) at 2-m height, which minimizes the ambiguity of measurements near of the surface. Specifically, is the temperature sensor measuring open water, snow, sea ice, or air? But now, with the dramatic decrease in Arctic SIE, increase in open water during summer, and the frailty of the younger sea ice pack, the IABP has had to deploy and develop new instruments to measure temperature. These instruments include Surface Velocity Program (SVP) buoys, which are commonly deployed on the world's ice-free oceans and typically measure sea surface temperature (SST), and the new robust Airborne eXpendable Ice Beacons (AXIB), which measure both SST and SAT. "Best Practice" requires that these instruments are inter-compared, and early results showing differences in collocated temperature measurements of over 2°C prompted the establishment of the IABP Arctic Observing Experiment (AOX) buoy test site at the US Department of Energy (DOE) Atmospheric Radiation Measurement (ARM) site in Barrow, Alaska. Preliminary results showed that the color of the hull of SVP buoys introduces a bias due to solar heating of the buoy. Since then, we have recommended that buoys should be painted white to reduce biases in temperature measurements due to different colors of the buoys deployed in different regions of the Arctic or the Antarctic. Measurements of SAT are more robust, but some of the temperature shields are susceptible to frosting. During our presentation we will provide an intercomparison of the temperature measurements at the AOX test site (i.e. high quality DOE/ARM observations compared with

  3. Simultaneous moduli measurement of elastic materials at elevated temperatures using an ultrasonic waveguide method

    Science.gov (United States)

    Periyannan, Suresh; Balasubramaniam, Krishnan

    2015-11-01

    A novel technique for simultaneously measuring the moduli of elastic isotropic material, as a function of temperature, using two ultrasonic guided wave modes that are co-generated using a single probe is presented here. This technique can be used for simultaneously measuring Young's modulus (E) and shear modulus (G) of different materials over a wide range of temperatures (35 °C-1200 °C). The specimens used in the experiments have special embodiments (for instance, a bend) at one end of the waveguide and an ultrasonic guided wave generator/detector (transducer) at the other end for obtaining reflected signals in a pulse-echo mode. The orientation of the transducer can be used for simultaneously generating/receiving the L(0,1) and/or T(0,1) using a single transducer in a waveguide on one end. The far end of the waveguides with the embodiment is kept inside a heating device such as a temperature-controlled furnace. The time of flight difference, as a function of uniform temperature distribution region (horizontal portion) of bend waveguides was measured and used to determine the material properties. Several materials were tested and the comparison between values reported in the literature and measured values were found to be in agreement, for both elastic moduli (E and G) measurements, as a function of temperature. This technique provides significant reduction in time and effort over conventional means of measurement of temperature dependence of elastic moduli.

  4. The systematic error of temperature noise correlation measurement method and self-calibration

    International Nuclear Information System (INIS)

    Tian Hong; Tong Yunxian

    1993-04-01

    The turbulent transport behavior of fluid noise and the nature of noise affect on the velocity measurement system have been studied. The systematic error of velocity measurement system is analyzed. A theoretical calibration method is proposed, which makes the velocity measurement of time-correlation as an absolute measurement method. The theoretical results are in good agreement with experiments

  5. Estimation of Surface Temperature and Heat Flux by Inverse Heat Transfer Methods Using Internal Temperatures Measured While Radiantly Heating a Carbon/Carbon Specimen up to 1920 F

    Science.gov (United States)

    Pizzo, Michelle; Daryabeigi, Kamran; Glass, David

    2015-01-01

    The ability to solve the heat conduction equation is needed when designing materials to be used on vehicles exposed to extremely high temperatures; e.g. vehicles used for atmospheric entry or hypersonic flight. When using test and flight data, computational methods such as finite difference schemes may be used to solve for both the direct heat conduction problem, i.e., solving between internal temperature measurements, and the inverse heat conduction problem, i.e., using the direct solution to march forward in space to the surface of the material to estimate both surface temperature and heat flux. The completed research first discusses the methods used in developing a computational code to solve both the direct and inverse heat transfer problems using one dimensional, centered, implicit finite volume schemes and one dimensional, centered, explicit space marching techniques. The developed code assumed the boundary conditions to be specified time varying temperatures and also considered temperature dependent thermal properties. The completed research then discusses the results of analyzing temperature data measured while radiantly heating a carbon/carbon specimen up to 1920 F. The temperature was measured using thermocouple (TC) plugs (small carbon/carbon material specimens) with four embedded TC plugs inserted into the larger carbon/carbon specimen. The purpose of analyzing the test data was to estimate the surface heat flux and temperature values from the internal temperature measurements using direct and inverse heat transfer methods, thus aiding in the thermal and structural design and analysis of high temperature vehicles.

  6. Temperature measurements in fluid flows (eventually reactive, multi-phase...) using optical methods; Mesure des temperatures dans les ecoulements (eventuellement reactifs, multiphasiques...) par methodes optiques

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    1998-12-31

    This conference day was organized by the `radiations` section of the French association of thermal engineers. This book of proceedings contains 8 papers entitled: `simultaneous temperature and NO concentration measurements in a hydrogen-air turbulent flame`; `application of iodine laser induced fluorescence to temperature, pressure and velocity measurements`; `Doppler phase measurement of refractive index and temperature`; `experimental and numerical study of temperature fields of particulates in plasma jets`; `measurement and determination of temperatures and concentrations of hot exhaust gases with FTIR emission spectroscopy`; `combustion control in gas turbines using CO{sub 2} emission spectroscopy`; `analysis of gases temperature fields and particulate jets. Application to hydrogen-air, kerosene stato-reactors and to solid propellant jets`; `restitution of temperature and species profiles in pre-mixing flames by inversion of transmission and IR emission data. (J.S.)

  7. Development of precise measurement method of neutron energy for plasma temperature diagnostics in thermonuclear fusion

    International Nuclear Information System (INIS)

    Mori, Chizuo; Gotoh, Junichi; Uritani, Akira; Miyahara, Hiroshi; Ikeda, Yuichiro; Kasugai, Yoshimi; Kaneko, Junichi

    1998-01-01

    There are many types of fast neutron spectrometers for plasma temperature diagnostics, 28 Si(n,α) 25 Mg reaction giving the energy resolution of 2.2% for 14 MeV neutrons, the 12 C(n,α) 9 Be reaction giving the resolution of 2.15%. These detectors, however suffer from radiation damage, which demands to exchange the detector to a new one in every a few month depending on the usage. Recoil proton method has also been developed by using liquid scintillator or plastic scintillator, as a neutron-to-proton converter in front of a Si-detector, which is called counter telescope type, giving a resolution of 4.0%. This type of spectrometer can reduce radiation damage by placing Si-detector at outside Neutron beam. The scintillator can measure the lost energy of protons in the converter (i.e. the scintillator) and the measured energy loss can be used for improving the energy resolution. However, the energy resolution of organic scintillator itself is generally not so good. We proposed to use a proportional counter with CH 4 as counting gas and also as a neutron-proton converter, which has far better energy resolution than plastic scintillators, although the time resolution of counting in proportional counters is generally inferior to that in organic scintillation counters. The characteristics of the new spectrometer were experimentally studied and also were simulated with analytical calculation. (author)

  8. Extending reversed-flow chromatographic methods for the measurement of diffusion coefficients to higher temperatures.

    Science.gov (United States)

    McGivern, W Sean; Manion, Jeffrey A

    2011-11-18

    A reversed-flow gas-chromatography (RF-GC) apparatus for the measurement of binary diffusion coefficients is described and utilized to measure the binary diffusion coefficients for several systems at temperatures from (300 to 723)K. Hydrocarbons are detected using flame ionization detection, and inert species can be detected by thermal conductivity. The present apparatus has been utilized to measure diffusion coefficients at substantially higher temperatures than previous RF-GC work. Characterization of the new apparatus was accomplished by comparing measured binary diffusion coefficients of dilute argon in helium to established reference values. Further diffusion coefficient measurements for dilute helium in argon and dilute nitrogen in helium (using thermal conductivity detection) and dilute methane in helium (using flame ionization detection) were performed and found to be in excellent agreement with literature values. The measurement of these well-established diffusion coefficients has shown that specific experimental conditions are required for accurate diffusion measurements using this technique, particularly at higher temperatures. Numerical simulations of the diffusion experiments are presented to demonstrate that artifacts of the analysis procedure must be specifically identified to ensure accuracy, particularly at higher temperatures. Published by Elsevier B.V.

  9. Method and Apparatus for measuring Gravitational Acceleration Utilizing a high Temperature Superconducting Bearing

    Energy Technology Data Exchange (ETDEWEB)

    Hull, John R.

    1998-11-06

    Gravitational acceleration is measured in all spatial dimensions with improved sensitivity by utilizing a high temperature superconducting (HTS) gravimeter. The HTS gravimeter is comprised of a permanent magnet suspended in a spaced relationship from a high temperature superconductor, and a cantilever having a mass at its free end is connected to the permanent magnet at its fixed end. The permanent magnet and superconductor combine to form a bearing platform with extremely low frictional losses, and the rotational displacement of the mass is measured to determine gravitational acceleration. Employing a high temperature superconductor component has the significant advantage of having an operative temperature at or below 77K, whereby cooling maybe accomplished with liquid nitrogen.

  10. Method and apparatus for measuring gravitational acceleration utilizing a high temperature superconducting bearing

    Science.gov (United States)

    Hull, John R.

    2000-01-01

    Gravitational acceleration is measured in all spatial dimensions with improved sensitivity by utilizing a high temperature superconducting (HTS) gravimeter. The HTS gravimeter is comprised of a permanent magnet suspended in a spaced relationship from a high temperature superconductor, and a cantilever having a mass at its free end is connected to the permanent magnet at its fixed end. The permanent magnet and superconductor combine to form a bearing platform with extremely low frictional losses, and the rotational displacement of the mass is measured to determine gravitational acceleration. Employing a high temperature superconductor component has the significant advantage of having an operating temperature at or below 77K, whereby cooling may be accomplished with liquid nitrogen.

  11. Methods for quantifying the influences of pressure and temperature variation on metal hydride reaction rates measured under isochoric conditions.

    Science.gov (United States)

    Voskuilen, Tyler G; Pourpoint, Timothée L

    2013-11-01

    Analysis techniques for determining gas-solid reaction rates from gas sorption measurements obtained under non-constant pressure and temperature conditions often neglect temporal variations in these quantities. Depending on the materials in question, this can lead to significant variations in the measured reaction rates. In this work, we present two new analysis techniques for comparison between various kinetic models and isochoric gas measurement data obtained under varying temperature and pressure conditions in a high pressure Sievert system. We introduce the integral pressure dependence method and the temperature dependence factor as means of correcting for experimental variations, improving model-measurement fidelity, and quantifying the effect that such variations can have on measured reaction rates. We use measurements of hydrogen absorption in LaNi5 and TiCrMn to demonstrate the effect of each of these methods and show that their use can provide quantitative improvements in interpretation of kinetics measurements.

  12. PHOTOEMISSION METHOD OF TEMPERATURE MEASURING IN THE PROCESS OF SPARK PLASMA SINTERING POWDERS OF REFRACTORY METALS

    Directory of Open Access Journals (Sweden)

    D. V. Minko

    2012-01-01

    Full Text Available Construction and algorithm of the photoemission pyrometer based on a photomultiplier are outlined; the calibration procedure is set out. The application of the photoemission method in a rapidly changing temperature is showed. It is proved that during spark plasma sintering the maximal temperature is 5500– 7500 °C, while the speed of temperature increase to its maximum ~108–109 °C/s, and the rate of decrease to 2000–4000 °C may be 106–107 °C/sec. It is recommended to use photoemission method when developing technological conditions, adjustment and control of technological processes using modern high-energy equipment (plasma, laser, cathode-ray tube to produce new materials, coatings and products for mechanical engineering, electronic industry and medicine.

  13. Electron Temperature Measurement in a Premixed Flat Flame Using the Double Probe Method

    Czech Academy of Sciences Publication Activity Database

    Wild, J.; Kudrna, P.; Tichý, M.; Nevrlý, V.; Střižík, M.; Bitala, P.; Filipi, B.; Zelinger, Zdeněk

    2012-01-01

    Roč. 52, č. 8 (2012), s. 692-698 ISSN 0863-1042 R&D Projects: GA ČR(CZ) GAP108/11/1312; GA MŠk LD12020 Institutional support: RVO:61388955 Keywords : double probe method * electron temperature * atmospheric premixed flame Subject RIV: CF - Physical ; Theoretical Chemistry Impact factor: 0.934, year: 2012

  14. Comparison of methods for estimating Wet-Bulb Globe Temperature index from standard meteorological measurements.

    Science.gov (United States)

    Patel, Tejash; Mullen, Stephen P; Santee, William R

    2013-08-01

    Environmental heat illness and injuries are a serious concern for the Army and Marines. Currently, the Wet-Bulb Globe Temperature (WBGT) index is used to evaluate heat injury risk. The index is a weighted average of dry-bulb temperature (Tdb), black globe temperature (Tbg), and natural wet-bulb temperature (Tnwb). The WBGT index would be more widely used if it could be determined using standard weather instruments. This study compares models developed by Liljegren at Argonne National Laboratory and by Matthew at the U.S. Army Institute of Environmental Medicine that calculate WBGT using standard meteorological measurements. Both models use air temperature (Ta), relative humidity, wind speed, and global solar radiation (RG) to calculate Tnwb and Tbg. The WBGT and meteorological data used for model validation were collected at Griffin, Georgia and Yuma Proving Ground (YPG), Arizona. Liljegren (YPG: R(2) = 0.709, p < 0.01; Griffin: R(2) = 0.854, p < 0.01) showed closer agreement between calculated and actual WBGT than Matthew (YPG: R(2) = 0.630, p < 0.01; Griffin: R(2) = 0.677, p < 0.01). Compared to actual WBGT heat categorization, the Matthew model tended to underpredict compared to Liljegren's classification. Results indicate Liljegren is an acceptable alternative to direct WBGT measurement, but verification under other environmental conditions is needed. Reprint & Copyright © 2013 Association of Military Surgeons of the U.S.

  15. Thermal diffusivity measurements of liquid materials at high temperature with the ''laser flash'' method

    International Nuclear Information System (INIS)

    Otter, Claude; Vandevelde, Jean

    1982-01-01

    Two solutions, one analytical and the other numerical are proposed to solve the thermokinetic problem encountered when measuring the thermal diffusivity of liquid materials at very high temperature (T>3123K). The liquid material is contained in a parallel faced vessel. This liquid is traversed by a short thermal pulse from a relaxed laser. The temperature response of the back face of the measurement cell is analysed. The first model proposed which does not take thermal losses into consideration, is a mathematical model derived from the ''two layer model'' (Larson and Koyama, 1968) extended to ''three layers''. In order to take the possibility of thermal losses to the external environment at high temperature into consideration, a Crank-Nicolson (1947) type numerical model utilizing finite differences is employed. These thermokinetic studies were performed in order to interpret temperature response curves obtained from the back face of a tungsten-liquid UO 2 -tungsten thermal wall, the purpose of the measurements made being to determine the thermal properties of liquid uranium oxide [fr

  16. Comparison of temperature measurement methods for evaluation of the thermal environment in vehicles

    DEFF Research Database (Denmark)

    Rosendahl, J..; Olesen, Bjarne W.

    2006-01-01

    A new standard, ISO/DIS 14505 is under preparation, dealing with the assessment of the thermal environment in vehicles, based on the equivalent temperature. The scope of this paper is to demonstrate the different results obtained when evaluating the thermal environment inside a vehicle, using the...... during each test. Based on the measurement results, the time used to reach the level of thermal comfort is evaluated and compared....

  17. A non-intrusive method for temperature measurements in flames produced by milligram-sized solid samples

    Science.gov (United States)

    Frances, Colleen Elizabeth

    Fires are responsible for the loss of thousands of lives and billions of dollars in property damage each year in the United States. Flame retardants can assist in the prevention of fires through mechanisms which either prevent or greatly inhibit flame spread and development. In this study samples of both brominated and non-brominated polystyrene were tested in the Milligram-scale Flaming Calorimeter and images captured with two DSL-R cameras were analyzed to determine flame temperatures through use of a non-intrusive method. Based on the flame temperature measurement results, a better understanding of the gas phase mechanisms of flame retardants may result, as temperature is an important diagnostic in the study of fire and combustion. Measurements taken at 70% of the total flame height resulted in average maximum temperatures of about 1656 K for polystyrene and about 1614 K for brominated polystyrene, suggesting that the polymer flame retardant may reduce flame temperatures.

  18. High temperature measuring device

    Science.gov (United States)

    Tokarz, Richard D.

    1983-01-01

    A temperature measuring device for very high design temperatures (to 2,000.degree. C.). The device comprises a homogenous base structure preferably in the form of a sphere or cylinder. The base structure contains a large number of individual walled cells. The base structure has a decreasing coefficient of elasticity within the temperature range being monitored. A predetermined quantity of inert gas is confined within each cell. The cells are dimensionally stable at the normal working temperature of the device. Increases in gaseous pressure within the cells will permanently deform the cell walls at temperatures within the high temperature range to be measured. Such deformation can be correlated to temperature by calibrating similarly constructed devices under known time and temperature conditions.

  19. An ultrasound-based liquid pressure measurement method in small diameter pipelines considering the installation and temperature.

    Science.gov (United States)

    Li, Xue; Song, Zhengxiang

    2015-04-09

    Liquid pressure is a key parameter for detecting and judging faults in hydraulic mechanisms, but traditional measurement methods have many deficiencies. An effective non-intrusive method using an ultrasound-based technique to measure liquid pressure in small diameter (less than 15 mm) pipelines is presented in this paper. The proposed method is based on the principle that the transmission speed of an ultrasonic wave in a Kneser liquid correlates with liquid pressure. Liquid pressure was calculated using the variation of ultrasonic propagation time in a liquid under different pressures: 0 Pa and X Pa. In this research the time difference was obtained by an electrical processing approach and was accurately measured to the nanosecond level through a high-resolution time measurement module. Because installation differences and liquid temperatures could influence the measurement accuracy, a special type of circuit called automatic gain control (AGC) circuit and a new back propagation network (BPN) model accounting for liquid temperature were employed to improve the measurement results. The corresponding pressure values were finally obtained by utilizing the relationship between time difference, transient temperature and liquid pressure. An experimental pressure measurement platform was built and the experimental results confirm that the proposed method has good measurement accuracy.

  20. Temperature measurement and control

    CERN Document Server

    Leigh, JR

    1988-01-01

    This book treats the theory and practice of temperature measurement and control and important related topics such as energy management and air pollution. There are no specific prerequisites for the book although a knowledge of elementary control theory could be useful. The first half of the book is an application oriented survey of temperature measurement techniques and devices. The second half is concerned mainly with temperature control in both simple and complex situations.

  1. Formaldehyde preparation methods for pressure and temperature dependent laser-induced fluorescence measurements

    Science.gov (United States)

    Burkert, A.; Müller, D.; Rieger, S.; Schmidl, G.; Triebel, W.; Paa, W.

    2015-12-01

    Formaldehyde is an excellent tracer for the early phase of ignition of hydrocarbon fuels and can be used, e.g., for characterization of single droplet ignition. However, due to its fast thermal decomposition at elevated temperatures and pressures, the determination of concentration fields from laser-induced fluorescence (LIF) measurements is difficult. In this paper, we address LIF measurements of this important combustion intermediate using a calibration cell. Here, formaldehyde is created from evaporation of paraformaldehyde. We discuss three setups for preparation of formaldehyde/air mixtures with respect to their usability for well-defined heating of formaldehyde/air mixtures. The "basic setup" uses a resist heater around the measurement cell for investigation of formaldehyde near vacuum conditions or formaldehyde/air samples after sequential admixing of air. The second setup, described for the first time in detail here, takes advantage of a constant flow formaldehyde/air regime which uses preheated air to reduce the necessary time for gas heating. We used the constant flow system to measure new pressure dependent LIF excitation spectra in the 343 nm spectral region (414 absorption band of formaldehyde). The third setup, based on a novel concept for fast gas heating via excitation of SF6 (chemically inert gas) using a TEA (transverse excitation at atmospheric pressure) CO2 laser, allows to further minimize both gas heating time and thermal decomposition. Here, an admixture of CO2 is served for real time temperature measurement based on Raman scattering. The applicability of the fast laser heating system has been demonstrated with gas mixtures of SF6 + air, SF6 + N2, as well as SF6 + N2 + CO2 at 1 bar total pressure.

  2. Influence of temperature on measurements of the CO2 compensation point: differences between the Laisk and O2-exchange methods

    Science.gov (United States)

    Walker, Berkley J.; Cousins, Asaph B.

    2013-01-01

    The CO2 compensation point in the absence of day respiration (Γ*) is a key parameter for modelling leaf CO2 exchange. Γ* links the kinetics of ribulose-1,5-bisphosphate carboxylase-oxygenase (Rubisco) with the stoichiometry of CO2 released per Rubisco oxygenation from photorespiration (α), two essential components of biochemical models of photosynthesis. There are two main gas-exchange methods for measuring Γ*: (i) the Laisk method, which requires estimates of mesophyll conductance to CO2 (g m) and (ii) measurements of O2 isotope exchange, which assume constant values of α and a fixed stoichiometry between O2 uptake and Rubisco oxygenation. In this study, the temperature response of Γ* measured using the Laisk and O2-exchange methods was compared under ambient (25 °C) and elevated (35 °C) temperatures to determine whether both methods yielded similar results. Previously published temperature responses of Γ* estimated with the Laisk and O2-exchange methods in Nicotiana tabacum demonstrated that the Laisk-derived model of Γ* was more sensitive to temperature compared with the O2-exchange model. Measurements in Arabidopsis thaliana indicated that the Laisk and O2-exchange methods produced similar Γ* at 25 °C; however, Γ* values from O2 exchange were lower at 35 °C compared with the Laisk method. Compared with a photorespiratory mutant (pmdh1pmdh2hpr) with increased α, wild-type (WT) plants had lower Laisk values of Γ* at 25 °C but were not significantly different at 35 °C. These differences between Laisk and O2 exchange values of Γ* at 35 °C could be explained by temperature sensitivity of α in WT and/or errors in the assumptions of O2 exchange. The differences between Γ* measured using the Laisk and O2-exchange method with temperature demonstrate that assumptions used to measure Γ*, and possibly the species-specific validity of these assumptions, need to be considered when modelling the temperature response of photosynthesis. PMID:23630324

  3. Neutron ion temperature measurement

    International Nuclear Information System (INIS)

    Strachan, J.D.; Hendel, H.W.; Lovberg, J.; Nieschmidt, E.B.

    1986-11-01

    One important use of fusion product diagnostics is in the determination of the deuterium ion temperature from the magnitude of the 2.5 MeV d(d,n) 3 He neutron emission. The detectors, calibration methods, and limitations of this technique are reviewed here with emphasis on procedures used at PPPL. In most tokamaks, the ion temperature deduced from neutrons is in reasonable agreement with the ion temperature deduced by other techniques

  4. A Hybrid Maximum Power Point Search Method Using Temperature Measurements in Partial Shading Conditions

    Directory of Open Access Journals (Sweden)

    Mroczka Janusz

    2014-12-01

    Full Text Available Photovoltaic panels have a non-linear current-voltage characteristics to produce the maximum power at only one point called the maximum power point. In the case of the uniform illumination a single solar panel shows only one maximum power, which is also the global maximum power point. In the case an irregularly illuminated photovoltaic panel many local maxima on the power-voltage curve can be observed and only one of them is the global maximum. The proposed algorithm detects whether a solar panel is in the uniform insolation conditions. Then an appropriate strategy of tracking the maximum power point is taken using a decision algorithm. The proposed method is simulated in the environment created by the authors, which allows to stimulate photovoltaic panels in real conditions of lighting, temperature and shading.

  5. Measurement of carbon dioxide flux from tropical peatland in Indonesia using the nocturnal temperature-inversion trap method

    Science.gov (United States)

    Iriana, Windy; Tonokura, Kenichi; Kawasaki, Masahiro; Inoue, Gen; Kusin, Kitso; Limin, Suwido H.

    2016-09-01

    Evaluation of CO2 flux from peatland soil respiration is important to understand the effect of land use change on the global carbon cycle and climate change and particularly to support carbon emission reduction policies. However, quantitative estimation of emitted CO2 fluxes in Indonesia is constrained by existing field data. Current methods for CO2 measurement are limited by high initial cost, manpower, and the difficulties associated with construction issues. Measurement campaigns were performed using a newly developed nocturnal temperature-inversion trap method, which measures the amount of CO2 trapped beneath the nocturnal inversion layer, in the dry season of 2013 at a drained tropical peatland near Palangkaraya, Central Kalimantan, Indonesia. This method is cost-effective and data processing is easier than other flux estimation methods. We compared CO2 fluxes measured using this method with the published data from the existing eddy covariance and closed chamber methods. The maximum value of our measurement results was 10% lower than maximum value of eddy covariance method and average value was 6% higher than average of chamber method in drained tropical peatlands. In addition, the measurement results shows good correlation with groundwater table. The results of this comparison suggest that this methodology for the CO2 flux measurement is useful for field research in tropical peatlands.

  6. Measurement of temperature profiles in process-applications using fibre-optical methods; Prozessgeeignete Temperaturprofilmessungen mit faseroptischen Methoden

    Energy Technology Data Exchange (ETDEWEB)

    Seefeld, P. [Endress und Hauser Wetzer GmbH und Co.KG, Nesselwang (Germany)

    2008-07-01

    Fibre-optical temperature measuring methods are offering an approach to detect temperature profiles. According to the NAMUR-Technology-Roadmap the detection of temperature profiles is representing an increased benefit. Intrinsic fibre-optical temperature measuring techniques are presented, known as OTDRmethod (Optical tine domain reflectometry) facilitating a distributed temperature measurement method that allows a resolution in the range of decimetres. For the purpose of such applications a suitable photoncounting device comprises mechanical robust fibre-optical components, 3 db Coupler, referenced Laser- Diode, Y-Coupler with integrated band-filter and APD (Avalanche Diode) used in a detection module. A VHDL-coded FBGA-board provides a basic control-device for - a Laser-Driver to generate adjustable exiting-pulses in the range of nanoseconds at rates up to 100 kHz. - a Photon-Counting module with a minimum opening width in the range of one nanosecond - that permits the co-addition of the photon-counts derived from the spectral resolved Stokes and Anti-Stokes band. (orig.)

  7. Temperature measurement device

    International Nuclear Information System (INIS)

    Oltman, B.G.; Eckerman, K.F.; Romberg, G.P.; Prepejchal, W.

    1975-01-01

    Thermoluminescent dosimeter (TLD) material is exposed to a known amount of radiation and then exposed to the environment where temperature measurements are to be taken. After a predetermined time period, the TLD material is read in a known manner to determine the amount of radiation energy remaining in the TLD material. The difference between the energy originally stored by irradiation and that remaining after exposure to the temperature ofthe environment is a measure of the average temperature of the environment during the exposure. (U.S.)

  8. A new lithium-ion battery internal temperature on-line estimate method based on electrochemical impedance spectroscopy measurement

    Science.gov (United States)

    Zhu, J. G.; Sun, Z. C.; Wei, X. Z.; Dai, H. F.

    2015-01-01

    The power battery thermal management problem in EV (electric vehicle) and HEV (hybrid electric vehicle) has been widely discussed, and EIS (electrochemical impedance spectroscopy) is an effective experimental method to test and estimate the status of the battery. Firstly, an electrochemical-based impedance matrix analysis for lithium-ion battery is developed to describe the impedance response of electrochemical impedance spectroscopy. Then a method, based on electrochemical impedance spectroscopy measurement, has been proposed to estimate the internal temperature of power lithium-ion battery by analyzing the phase shift and magnitude of impedance at different ambient temperatures. Respectively, the SoC (state of charge) and temperature have different effects on the impedance characteristics of battery at various frequency ranges in the electrochemical impedance spectroscopy experimental study. Also the impedance spectrum affected by SoH (state of health) is discussed in the paper preliminary. Therefore, the excitation frequency selected to estimate the inner temperature is in the frequency range which is significantly influenced by temperature without the SoC and SoH. The intrinsic relationship between the phase shift and temperature is established under the chosen excitation frequency. And the magnitude of impedance related to temperature is studied in the paper. In practical applications, through obtaining the phase shift and magnitude of impedance, the inner temperature estimation could be achieved. Then the verification experiments are conduced to validate the estimate method. Finally, an estimate strategy and an on-line estimation system implementation scheme utilizing battery management system are presented to describe the engineering value.

  9. An investigation of methods for neutron dose measurement in high temperature irradiation fields

    Energy Technology Data Exchange (ETDEWEB)

    Kosako, Toshisou; Sugiura, Nobuyuki [Tokyo Univ. (Japan); Kudo, Kazuhiko [Kyushu Univ., Fukuoka (Japan)] [and others

    2000-10-01

    The Japan Atomic Energy Research Institute (JAERI) has been conducting the innovative basic research on high temperature since 1994, which is a series of high temperature irradiation studies using the High Temperature Engineering Test Reactor (HTTR). 'The Task Group for Evaluation of Irradiation Dose under High Temperature Radiation' was founded in the HTTR Utilization Research Committee, which is the promoting body of the innovative basic research. The present report is a summary of investigation which has been made by the Task Group on the present status and subjects of research and development of neutron detectors in high temperature irradiation fields, in view of contributing to high temperature irradiation research using the HTTR. Detectors investigated here in the domestic survey are the following five kinds of in-core detectors: 1) small fission counter, 2) small fission chamber, 3) self-powered detector, 4) activation detector, and 5) optical fiber. In addition, the research and development status in Russia has been investigated. The present report will also be useful as nuclear instrumentation of high temperature gas-cooled reactors. (author)

  10. Feasibility Studies of the Two Filters Method in TJ-II for Electron Temperature Measurements in High Density Plasmas

    International Nuclear Information System (INIS)

    Baiao, D.; Medina, F.; Ochando, M.; Varandas, C.

    2009-01-01

    The TJ-II plasma soft X-ray emission was studied in order to establish an adequate setup for an electron temperature diagnostic suitable for high density, with spatial and temporal resolutions, based on the two-filters method. The preliminary experimental results reported were obtained with two diagnostics (an X-ray PHA based on a Ge detector and a tomography system) already installed in TJ-II stellarator. These results lead to the conclusion that the two-filters method was a suitable option for an electron temperature diagnostic for high-density plasmas in TJ-II. We present the design and fi rst results obtained with a prototype for the measurement of electron temperature in TJ-II plasmas heated with energetic neutral beams. This system consists in two AXUV20A detectors which measure the soft X-ray plasma emissivity trough beryllium filters of different thickness. From the two-filters technique it is possible to estimate the electron temperature. The analyses carried out allowed concluding which filter thicknesses are most suited for TJ-II plasmas, and enhanced the need of a computer code to simulate signals and plasma compositions. (Author) 7 refs.

  11. Feasibility Studies of the Two Filters Method in TJ-II for Electron Temperature Measurements in High Density Plasmas

    Energy Technology Data Exchange (ETDEWEB)

    Baiao, D.; Medina, F.; Ochando, M.; Varandas, C.

    2009-07-01

    The TJ-II plasma soft X-ray emission was studied in order to establish an adequate setup for an electron temperature diagnostic suitable for high density, with spatial and temporal resolutions, based on the two-filters method. The preliminary experimental results reported were obtained with two diagnostics (an X-ray PHA based on a Ge detector and a tomography system) already installed in TJ-II stellarator. These results lead to the conclusion that the two-filters method was a suitable option for an electron temperature diagnostic for high-density plasmas in TJ-II. We present the design and fi rst results obtained with a prototype for the measurement of electron temperature in TJ-II plasmas heated with energetic neutral beams. This system consists in two AXUV20A detectors which measure the soft X-ray plasma emissivity trough beryllium filters of different thickness. From the two-filters technique it is possible to estimate the electron temperature. The analyses carried out allowed concluding which filter thicknesses are most suited for TJ-II plasmas, and enhanced the need of a computer code to simulate signals and plasma compositions. (Author) 7 refs.

  12. Temperature measurements by thermocouples

    International Nuclear Information System (INIS)

    Liermann, J.

    1975-01-01

    The measurement of a temperature (whatever the type of transducer used) raises three problems: the choice of transducer; where it should be placed; how it should be fixed and protected. These are the three main points examined, after a brief description of the most commonly used thermocouples [fr

  13. Measuring ac-loss in high temperature superconducting cable-conductors using four probe methods

    DEFF Research Database (Denmark)

    Kühle (fratrådt), Anders Van Der Aa; Træholt, Chresten; Olsen, Søren Krüger

    1999-01-01

    Measuring the ac-loss of superconducting cable conductors have many aspects in common with measuring the ac-loss of single superconducting tapes. In a cable conductor all tapes are connected to each other and to the test circuit through normal metal joints in each end. This makes such measurement...

  14. Ultrasonic computerized tomography (CT) for temperature measurements with limited projection data based on extrapolated filtered back projection (FBP) method

    International Nuclear Information System (INIS)

    Zhu Ning; Jiang Yong; Kato, Seizo

    2005-01-01

    This study uses ultrasound in combination with tomography to obtain three-dimensional temperature measurements using projection data obtained from limited projection angle. The main feature of the new computerized tomography (CT) reconstruction algorithm is to employ extrapolation scheme to make up for the incomplete projection data, it is based on the conventional filtered back projection (FBP) method while on top of that taking into account the correlation between the projection data and Fourier transform-based extrapolation. Computer simulation is conducted to verify the above algorithm. An experimental 3D temperature distribution measurement is also carried out to validate the proposed algorithm. The simulation and experimental results demonstrate that the extrapolated FBP CT algorithm is highly effective in dealing with projection data from limited projection angle

  15. Method of correction of measured temperature and velocity field in free hot gas jet

    Czech Academy of Sciences Publication Activity Database

    Gregor, J.; Jakubová, I.; Mendl, T.; Šenk, J.; Kopecký, Vladimír

    2002-01-01

    Roč. 52, supplement D (2002), s. 601-606 ISSN 0011-4626. [Symposium on Plasma Physics and Technology/20th./. Prague, 10.06.2002-13.06.2002] R&D Projects: GA ČR GA202/01/1563 Institutional research plan: CEZ:AV0Z2043910 Keywords : plasma, temperature profile, thermocouple Subject RIV: BL - Plasma and Gas Discharge Physics Impact factor: 0.311, year: 2002

  16. A Physically Motivated and Empirically Calibrated Method to Measure the Effective Temperature, Metallicity, and Ti Abundance of M Dwarfs

    Science.gov (United States)

    Veyette, Mark J.; Muirhead, Philip S.; Mann, Andrew W.; Brewer, John M.; Allard, France; Homeier, Derek

    2017-12-01

    The ability to perform detailed chemical analysis of Sun-like F-, G-, and K-type stars is a powerful tool with many applications, including studying the chemical evolution of the Galaxy and constraining planet formation theories. Unfortunately, complications in modeling cooler stellar atmospheres hinders similar analyses of M dwarf stars. Empirically calibrated methods to measure M dwarf metallicity from moderate-resolution spectra are currently limited to measuring overall metallicity and rely on astrophysical abundance correlations in stellar populations. We present a new, empirical calibration of synthetic M dwarf spectra that can be used to infer effective temperature, Fe abundance, and Ti abundance. We obtained high-resolution (R ˜ 25,000), Y-band (˜1 μm) spectra of 29 M dwarfs with NIRSPEC on Keck II. Using the PHOENIX stellar atmosphere modeling code (version 15.5), we generated a grid of synthetic spectra covering a range of temperatures, metallicities, and alpha-enhancements. From our observed and synthetic spectra, we measured the equivalent widths of multiple Fe I and Ti I lines and a temperature-sensitive index based on the FeH band head. We used abundances measured from widely separated solar-type companions to empirically calibrate transformations to the observed indices and equivalent widths that force agreement with the models. Our calibration achieves precisions in T eff, [Fe/H], and [Ti/Fe] of 60 K, 0.1 dex, and 0.05 dex, respectively, and is calibrated for 3200 K H] < +0.3, and -0.05 < [Ti/Fe] < +0.3. This work is a step toward detailed chemical analysis of M dwarfs at a precision similar to what has been achieved for FGK stars.

  17. Method for measuring temperatures and densities in hypersonic wind tunnel air flows using laser-induced O2 fluorescence

    Science.gov (United States)

    Laufer, Gabriel; Mckenzie, Robert L.; Fletcher, Douglas G.

    1990-01-01

    Laser-induced fluorescence in oxygen, in combination with Raman scattering, is shown to be an accurate means by which temperature, density, and their fluctuations owing to turbulence can be measured in air flows associated with high-speed wind tunnels. For temperatures above 60 K and densities above 0.01 amagat, the uncertainties in the temperature and density measurements can be less than 2 percent, if the signal uncertainties are dominated by photon statistical noise. The measurements are unaffected by collisional quenching and can be achieved with laser fluences for which nonlinear effects are insignificant. Temperature measurements using laser-induced fluorescence alone have been demonstrated at known densities in the range of low temperatures and densities which are expected in a hypersonic wind tunnel.

  18. A method for measuring temperatures and densities in hypersonic wind tunnel air flows using laser-induced O2 fluorescence

    Science.gov (United States)

    Laufer, Gabriel; Fletcher, Douglas G.; Mckenzie, Robert L.

    1990-01-01

    Laser-induced fluorescence in oxygen, in combination with Raman scattering, is shown to be an accurate means by which temperature, density, and their fluctuations due to turbulence can be measured in air flows associated with high-speed wind tunnels. For temperatures above 60 K and densities above 0.01 amagat, the uncertainty in the temperature and density measurements can be less than 2 and 3 percent, respectively, if the signal uncertainties are dominated by photon-statistical noise. The measurements are unaffected by collisional quenching and can be achieved with laser fluences for which nonlinear effects are insignificant. Temperature measurements using laser-induced fluorescence alone have been demonstrated at known densities in the range of low temperatures and densities which are expected in a hypersonic wind tunnel.

  19. Surface Temperature Measurement Using Hematite Coating

    Science.gov (United States)

    Bencic, Timothy J. (Inventor)

    2015-01-01

    Systems and methods that are capable of measuring temperature via spectrophotometry principles are discussed herein. These systems and methods are based on the temperature dependence of the reflection spectrum of hematite. Light reflected from these sensors can be measured to determine a temperature, based on changes in the reflection spectrum discussed herein.

  20. New algorithm for extreme temperature measurements

    NARCIS (Netherlands)

    Damean, N.

    2000-01-01

    A new algorithm for measurement of extreme temperature is presented. This algorithm reduces the measurement of the unknown temperature to the solving of an optimal control problem, using a numerical computer. Based on this method, a new device for extreme temperature measurements is projected. It

  1. Measurement of Internal Friction for Tungsten by the Curve Vibrating Method with Variation of Voltage and Temperature

    Directory of Open Access Journals (Sweden)

    Elin Yusibani

    2013-12-01

    Full Text Available Application of a curved vibrating wire method (CVM to measure gas viscosity has been widely used. A fine Tungsten wire with 50 mm of diameter is bent into a semi-circular shape and arranged symmetrically in a magnetic field of about 0.2 T. The frequency domain is used for calculating the viscosity as a response for forced oscillation of the wire. Internal friction is one of the parameter in the CVM which is has to be measured beforeahead. Internal friction coefficien for the wire material which is the inverse of the quality factor has to be measured in a vacuum condition. The term involving internal friction actually represents the effective resistance of motion due to all non-viscous damping phenomena including internal friction and magnetic damping. The testing of internal friction measurement shows that at different induced voltage and elevated temperature at a vacuum condition, it gives the value of internal friction for Tungsten is around 1 to 4 10-4.

  2. Radiometric temperature measurements fundamentals

    CERN Document Server

    Zhang, Zhuomin M; Machin, Graham

    2009-01-01

    This book describes the theory of radiation thermometry, both at a primary level and for a variety of applications, such as in the materials processing industries and remote sensing. This book is written for those who will apply radiation thermometry in industrial practice; use radiation thermometers for scientific research; the radiation thermometry specialist in a national measurement institute; developers of radiation thermometers who are working to innovate products for instrument manufacturers, and developers of non-contact thermometry methods to address challenging thermometry problems.

  3. An online Vce measurement and temperature estimation method for high power IGBT module in normal PWM operation

    DEFF Research Database (Denmark)

    Ghimire, Pramod; de Vega, Angel Ruiz; Beczkowski, Szymon

    2014-01-01

    An on-state collector-emitter voltage (Vce) measurement and thereby an estimation of average temperature in space for high power IGBT module is presented while power converter is in operation. The proposed measurement circuit is able to measure both high and low side IGBT and anti parallel diode...... voltages for a half bridge module which are also used to monitor the electrical degradation of the module. The Vce load current is proposed to estimate the variation of average temperature in space at every fundamental cycle of sinusoidal loading current. Initially, the calibration of voltage and junction...

  4. A survey of temperature measurement

    International Nuclear Information System (INIS)

    Saltvold, J.R.

    1976-03-01

    Many different techniques for measuring temperature have been surveyed and are discussed. The concept of temperature and the physical phenomena used in temperature measurement are also discussed. Extensive tables are presented in which the range and accuracy of the various techniques and other related data are included. (author)

  5. Observation of soft x-ray radiation from Heliotron E plasmas by the absorption method for the measurement of electron temperatures

    International Nuclear Information System (INIS)

    Kaneko, H.; Tohda, T.; Iiyoshi, A.

    1989-01-01

    An absorption method of soft x ray is applied to Heliotron E plasmas for measurement of the electron temperature. Nitrogen gas is used as an absorber for convenience, owing to its accurate, uniform, and easily controllable density. The general feature of the absorption method for measurement of the electron temperature is discussed using a model with two parameters: the generalized thickness of the absorber and the electron temperature. The energy resolution of this method is not sufficient as a general method for spectral analysis. Hence, it is necessary to assume in advance such a model spectrum as consists of bremsstrahlung, recombination radiation, and impurity line radiation. Since the spectrum is always assumed before the analysis, we should try to find the origins of deformation of the energy spectrum and to correct the contribution. The effect of line emission from impurity ions to the estimated electron temperature is evaluated as a function of the electron temperature and the energy of the line relative to the generalized absorber thickness used in the measurement. An actual spectrum is measured by a pulse-height analysis (PHA) of the soft x ray. The one clear line, from chlorine, is not significant in the present determination of the electron temperature by the absorption method. Another possible line from iron at energy less than 1 keV is included in the analysis. Using a convenient method for determination of local emissivity from a chord-integrated emissivity, the electron temperature is determined from the local emissivity. The observed broad electron-temperature profile might be an artifact due to recombination radiation of the highly ionized ion diffused out of the hot core of the plasma. It is confirmed that the absorption method gives absolute measurement of the electron temperature at the plasma center, when additional information on impurity lines are given by PHA

  6. Thermal transient effect and improved junction temperature measurement method in high-voltage light-emitting diodes

    NARCIS (Netherlands)

    Ye, H.; Chen, X.; Zeijl, H. van; Gielen, A.W.J.; Zhang, G.

    2013-01-01

    The diode forward voltage method with pulsed currents was widely used for monitoring junction temperature (Tj of light-emitting diodes (LEDs). However, a thermal transient effect (TTE) was observed by the pulsed currents and consequent errors were introduced. Thermoelectric physics was conducted to

  7. A Simple Method to Measure the Thermal contraction Percentage of a Solid Between Room and Liquid Nitrogen Temperatures

    International Nuclear Information System (INIS)

    Grau Carles, A.; Grau Malonda, A.

    2000-01-01

    We described how to build a simple device for measuring, with a reasonable good accuracy, the thermal contraction of a flat sample between room and liquid nitrogen temperatures. The contraction percentage of the sample is determined by the dimensional comparison of two images taken through the bottom of a transparent quartz tray. Instead of a photo or video camera, a high-resolution flatbed scanner is utilized to avoid the correction of perspectives. The so-called Grueneisen approximation are applied to evaluate the contraction percentages for intermediate temperatures. (Author) 28 refs

  8. A Simple Method to Measure the Thermal contraction Percentage of a Solid Between Room and Liquid Nitrogen Temperatures

    Energy Technology Data Exchange (ETDEWEB)

    Grau Carles, A.; Grau Malonda, A. [Ciemat, Madrid (Spain)

    2000-07-01

    We described how to build a simple device for measuring, with a reasonable good accuracy, the thermal contraction of a flat sample between room and liquid nitrogen temperatures. The contraction percentage of the sample is determined by the dimensional comparison of two images taken through the bottom of a transparent quartz tray. Instead of a photo or video camera, a high-resolution flatbed scanner is utilized to avoid the correction of perspectives. The so-called Grueneisen approximation are applied to evaluate the contraction percentages for intermediate temperatures. (Author) 28 refs.

  9. A method for the retrieval of atomic oxygen density and temperature profiles from ground-based measurements of the O(+)(2D-2P) 7320 A twilight airglow

    Science.gov (United States)

    Fennelly, J. A.; Torr, D. G.; Richards, P. G.; Torr, M. R.; Sharp, W. E.

    1991-01-01

    This paper describes a technique for extracting thermospheric profiles of the atomic-oxygen density and temperature, using ground-based measurements of the O(+)(2D-2P) doublet at 7320 and 7330 A in the twilight airglow. In this method, a local photochemical model is used to calculate the 7320-A intensity; the method also utilizes an iterative inversion procedure based on the Levenberg-Marquardt method described by Press et al. (1986). The results demonstrate that, if the measurements are only limited by errors due to Poisson noise, the altitude profiles of neutral temperature and atomic oxygen concentration can be determined accurately using currently available spectrometers.

  10. Field of Temperature Measurement by Virtual Instrumentation

    Directory of Open Access Journals (Sweden)

    Libor HARGAŠ

    2009-01-01

    Full Text Available This paper introduces about temperature determination for given dot of picture through image analysis. Heat transfer is the transition of thermal energy from a heated item to a cooler item. Main method of measurement of temperature in image is Pattern Matching, color scale detection and model detection. We can measure temperature dependency at time for selected point of thermo vision images. This measurement gives idea about the heat transfer at time dependences.

  11. Multichord time-resolved electron temperature measurements by the x-ray absorber-foil method on TFTR

    International Nuclear Information System (INIS)

    Kiraly, J.; Bitter, M.; Efthimion, P.

    1985-09-01

    Absorber foils have been installed in the TFTR X-Ray Imaging System to permit measurement of the electron temperature along 10 to 30 chords spaced at 5-12.5 cm with a time resolution of less than 100 μs. The technique uses the ratio of x-ray fluxes transmitted through two different foils. The ratio depends mainly on electron temperature. Simulations show that strong impurity line radiation can distort this ratio. To correct for these effects, special beryllium-scandium filters are employed to select the line-free region between 2 and 4.5 keV. Other filter pairs allow corrections for Fe L and Ni L line radiation as well as Ti K and Ni K emission. Good accuracy is also obtained with simple beryllium filters, provided that impurity corrections are incorporated in the analysis, taking line intensities from the x-ray pulse-height analysis diagnostic. A description of modeling calculations and a comparison of temperature values from this diagnostic with data from the x-ray pulse height analysis, the electron cyclotron emission, and the Thomson scattering diagnostics are presented. Several applications of the absorber foil electron temperature diagnostic on TFTR are discussed

  12. Temperature standards, what and where: resources for effective temperature measurements

    International Nuclear Information System (INIS)

    Johnston, W.W. Jr.

    1982-01-01

    Many standards have been published to describe devices, methods, and other topics. How they are developed and by whom are briefly described, and an attempt is made to extract most of those relating to temperature measurements. A directory of temperature standards and their sources is provided

  13. Safe and consistent method of spot-welding platinum thermocouple wires and foils for high temperature measurements

    Science.gov (United States)

    Orr, G.; Roth, M.

    2012-08-01

    A low-voltage (mV) electronically triggered spot welding system for fabricating fine thermocouples and thin sheets used in high-temperature characterization of materials' properties is suggested. The system is based on the capacitance discharge method with a timed trigger for obtaining reliable and consistent welds. In contrast to existing techniques based on employing high voltage DC supplies for charging the capacitor or supplies with positive and negative rails, this method uses a simple, standard dual power supply available at most of the physical laboratories or can be acquired at a low cost. In addition, an efficient and simple method of fabricating non-sticking electrodes that do not contaminate the weld area is suggested and implemented.

  14. A novel method to measure self-association of small amphipathic molecules: temperature profiling in reversed-phase chromatography.

    Science.gov (United States)

    Lee, Darin L; Mant, Colin T; Hodges, Robert S

    2003-06-20

    Biophysical techniques such as size-exclusion chromatography, sedimentation equilibrium analytical ultracentrifugation, and non-denaturing gel electrophoresis are the classical methods for determining the self-association of molecules into dimers, trimers, or other higher order species. However, these techniques usually require high (mg/ml) loading concentrations to detect self-association and also possess a lower size limit that is dependent on the ability of the technique to resolve monomeric from higher order species. Here we describe a novel, sensitive method with no upper or lower molecular size limits that indicates self-association of molecules driven together by the hydrophobic effect under aqueous conditions. "Temperature profiling in reversed-phase chromatography" analyzes the retention behavior of a sample over the temperature range of 5-80 degrees C during gradient elution reversed-phase high-performance liquid chromatography. Because this technique greatly increases the effective concentration of analyte upon adsorption to the column, it is extremely sensitive, requiring very small sample quantities (microgram or less). In contrast, the classical techniques mentioned above decrease the effective analyte concentration during analysis, decreasing sensitivity by requiring larger amounts of analyte to detect molecular self-association. We demonstrate the utility of this technique with 14-residue cyclic and linear cationic peptides (molecule when using this technique are its ability to be retained on the reversed-phase column and to be subsequently removed from the column during gradient elution.

  15. Temperature measurement in the sea

    Digital Repository Service at National Institute of Oceanography (India)

    Krishnamacharyulu, R.J.; Rao, L.V.G.

    ), their advantages and limitations are also touched upon. Calibration of various instruments used for temperature measurement in the sea and the special setup/facilities needed for this purpose are also discussed...

  16. Rate and Temperature Dependence of Adhesion Measured by a Jkr Method on Synthetically Modified Acrylic Pressure Sensitive Adhesives

    Science.gov (United States)

    Garif, Yev; Gerberich, William; Macosko, Christopher; Pocius, Alphonsus

    2003-03-01

    The goal of this study is to model mechanisms of interfacial separation in soft polymers based on experimental results obtained from normal mechanical contacts (JKR test). For the JKR test, cylindrically shaped samples of acrylic pressure sensitive adhesives (PSAs) were synthesized in capillary tubes in presence of a cross-linking agent in order to obtain an elastic PSA-like network (PSA-LN). Surface characteristics of individual samples were altered by co-polymerizing small amounts of polar side-groups such as acrylic acid (AA), dimethylaminoethylacrylate (DMAEA), and acrylonitrile (AN). The measurement was conducted in a ventilated chamber at three different temperatures (0, 25, and 50 degrees Celcius) and zero humidity with a sub-micron-per-second range of contact rates. Within this range, measured adhesion exhibits a transition from nearly rate-independent behavior to power law behavior at higher rates. Power law index matches that from the peel test data of similarly synthesized adhesive tapes. Accordingly, the transition is linked to the characteristic length scale of the process zone at the interface, as opposed to the bulk, ahead of the slowly advancing crack tip.

  17. Designing an accurate system for temperature measurements

    Directory of Open Access Journals (Sweden)

    Kochan Orest

    2017-01-01

    Full Text Available The method of compensation of changes in temperature field along the legs of inhomogeneous thermocouple, which measures a temperature of an object, is considered in this paper. This compensation is achieved by stabilization of the temperature field along the thermocouple. Such stabilization does not allow the error due to acquired thermoelectric inhomogeneity to manifest itself. There is also proposed the design of the furnace to stabilize temperature field along the legs of the thermocouple which measures the temperature of an object. This furnace is not integrated with the thermocouple mentioned above, therefore it is possible to replace this thermocouple with a new one when it get its legs considerably inhomogeneous.. There is designed the two loop measuring system with the ability of error correction which can use simultaneously a usual thermocouple as well as a thermocouple with controlled profile of temperature field. The latter can be used as a reference sensor for the former.

  18. Michelson interferometer for measuring temperature

    OpenAIRE

    Xie, Dong; Xu, Chunling; wang, Anmin

    2016-01-01

    We investigate that temperature can be measured by a modified Michelson interferometer, where at least one reflected mirror is replaced by a thermalized sample. Both of two mirrors replaced by the corresponding two thermalized samples can help to approximatively improve the resolution of temperature up to twice than only one mirror replaced by a thermalized sample. For further improving the precision, a nonlinear medium can be employed. The Michelson interferometer is embedded in a gas displa...

  19. Michelson interferometer for measuring temperature

    Science.gov (United States)

    Xie, Dong; Xu, Chunling; Wang, An Min

    2017-09-01

    We investigate that temperature can be measured by a modified Michelson interferometer, where at least one reflected mirror is replaced by a thermalized sample. Both of two mirrors replaced by the corresponding two thermalized samples can help to approximatively improve the resolution of temperature up to twice than only one mirror replaced by a thermalized sample. For further improving the precision, a nonlinear medium can be employed. The Michelson interferometer is embedded in a gas displaying Kerr nonlinearity. We obtain the analytical equations and numerically calculate the precision with parameters within the reach of current technology, proving that the precision of temperature can be greatly enhanced by using a nonlinear medium. Our results show that one can create an accurate thermometer by measuring the photons in the Michelson interferometer, with no need to directly measure the population of thermalized sample.

  20. A complexity measure based method for studying the dependance of 222Rn concentration time series on indoor air temperature and humidity.

    Science.gov (United States)

    Mihailovic, D T; Udovičić, V; Krmar, M; Arsenić, I

    2014-02-01

    We have suggested a complexity measure based method for studying the dependence of measured (222)Rn concentration time series on indoor air temperature and humidity. This method is based on the Kolmogorov complexity (KL). We have introduced (i) the sequence of the KL, (ii) the Kolmogorov complexity highest value in the sequence (KLM) and (iii) the KL of the product of time series. The noticed loss of the KLM complexity of (222)Rn concentration time series can be attributed to the indoor air humidity that keeps the radon daughters in air. © 2013 Published by Elsevier Ltd.

  1. Quantum interferometric measurements of temperature

    Science.gov (United States)

    Jarzyna, Marcin; Zwierz, Marcin

    2015-09-01

    We provide a detailed description of the quantum interferometric thermometer, which is a device that estimates the temperature of a sample from the measurements of the optical phase. We rigorously analyze the operation of such a device by studying the interaction of the optical probe system prepared in a single-mode Gaussian state with a heated sample modeled as a dissipative thermal reservoir. We find that this approach to thermometry is capable of measuring the temperature of a sample in the nanokelvin regime. Furthermore, we compare the fundamental precision of quantum interferometric thermometers with the theoretical precision offered by the classical idealized pyrometers, which infer the temperature from a measurement of the total thermal radiation emitted by the sample. We find that the interferometric thermometer provides a superior performance in temperature sensing even when compared with this idealized pyrometer. We predict that interferometric thermometers will prove useful for ultraprecise temperature sensing and stabilization of quantum optical experiments based on the nonlinear crystals and atomic vapors.

  2. Measuring nanowire thermal conductivity at high temperatures

    Science.gov (United States)

    Wang, Xiaomeng; Yang, Juekuan; Xiong, Yucheng; Huang, Baoling; Xu, Terry T.; Li, Deyu; Xu, Dongyan

    2018-02-01

    This work extends the micro-thermal-bridge method for thermal conductivity measurements of nanowires to high temperatures. The thermal-bridge method, based on a microfabricated device with two side-by-side suspended membranes with integrated platinum resistance heaters/thermometers, has been used to determine thermal conductivity of various nanowires/nanotubes/nanoribbons at relatively low temperatures. However, to date, thermal conductivity characterization of nanowires at temperatures above 600 K has seldom been reported presumably due to several technical difficulties including the instability of the microfabricated thermometers, radiation heat loss, and the effect of the background conductance on the measurement. Here we report on our attempt to address the aforementioned challenges and demonstrate thermal conductivity measurement of boron nanoribbons up to 740 K. To eliminate high temperature resistance instability, the device is first annealed at 1023 K for 5 min in an argon atmosphere. Two radiation shields are installed in the measurement chamber to minimize radiation heat loss from the measurement device to the surroundings; and the temperature of the device at each set point is calibrated by an additional thermocouple directly mounted on the chip carrier. The effect of the background conductance is eliminated by adopting a differential measurement scheme. With all these modifications, we successfully measured the thermal conductivity of boron nanoribbons over a wide temperature range from 27 K to 740 K. The measured thermal conductivity increases monotonically with temperature and reaches a plateau of ~2.5 W m‑1 K‑1 at approximately 400 K, with no clear signature of Umklapp scattering observed in the whole measurement temperature range.

  3. Saturation effects in Na lidar temperature measurements

    International Nuclear Information System (INIS)

    von der Gathen, P.

    1991-01-01

    Na atoms residing in the 80-110 km altitude region can be used to probe the Doppler-broadened hyperfine structure (hfs) of their D 2 resonance by ground-bases lidar and hence to deduce atmospheric temperatures. In principle, two different methods may be employed: (1) wavelength scanning of the hfs with a narrow-band laser and signal detection with a broad-band receiver, and (2) use of a broad-band laser and wavelength scanning of the hfs with a narrow-band receiver. These two methods are affected in different ways by laser-induced saturation in the Na layer, the effect on the measurements of sodium densities and of atmospheric temperatures being quite different. Density measurements are affected by the absolute level of saturation. Temperature measurements, however, are affected by the difference in saturation at the scanned wavelengths. If, additionally, observed signal levels are taken into account, method 1 is more efficient than method 2 for both types of measurements at nighttime, whereas a modified method 2 surpasses method 1 for temperature measurements at daytime

  4. Below-ground process responses to elevated CO2 and temperature: a discussion of observations, measurement methods, and models

    Science.gov (United States)

    Elise Pendall; Scott Bridgham; Paul J. Hanson; Bruce Hungate; David W. Kicklighter; Dale W. Johnson; Beverly E. Law; Yiqi Luo; J. Patrick Megonigal; Maria Olsrud; Michael G. Ryan; Shiqiang Wan

    2004-01-01

    Rising atmospheric CO2 and temperatures are probably altering ecosystem carbon cycling, causing both positive and negative feedbacks to climate. Below-ground processes play a key role in the global carbon (C) cycle because they regulate storage of large quantities of C, and are potentially very sensitive to direct and indirect effects of elevated...

  5. A new method to measure bowen ratios using high resolution vertical dry and wet bulb temperature profiles

    NARCIS (Netherlands)

    Euser, T.; Luxemburg, W.; Everson, C.; Mengistu, M.; Clulow, A.; Bastiaanssen, W.

    2013-01-01

    The Bowen ratio surface energy balance method is a relatively simple method to determine the latent heat flux and the actual land surface evaporation. Despite its simplicity, the Bowen ratio method is generally considered to be unreliable due to the use of two-level sensors that are installed by

  6. Noise-Measuring Method

    DEFF Research Database (Denmark)

    Diamond, J. M.

    1965-01-01

    A noise-measuring method based on the use of a calibrated noise generator and an output meter with a special scale is described. The method eliminates the effect of noise contributions occurring in the circuits following the device under test.......A noise-measuring method based on the use of a calibrated noise generator and an output meter with a special scale is described. The method eliminates the effect of noise contributions occurring in the circuits following the device under test....

  7. Experimental Evaluation of IGBT Junction Temperature Measurement via a Modified-VCE (ΔVCE_ΔVGE) Method with Series Resistance Removal

    DEFF Research Database (Denmark)

    Baker, Nick; Iannuzzo, Francesco; Munk-Nielsen, Stig

    2016-01-01

    Temperature sensitive electrical parameters (TSEPs) allow junction temperature measurements on power semiconductors without modification to module packaging. The on-state collector-emitter voltage (VCE) in IGBTs is an attractive option for junction temperature measurement since it is relatively e...

  8. Cutting temperature measurement and material machinability

    Directory of Open Access Journals (Sweden)

    Nedić Bogdan P.

    2014-01-01

    Full Text Available Cutting temperature is very important parameter of cutting process. Around 90% of heat generated during cutting process is then away by sawdust, and the rest is transferred to the tool and workpiece. In this research cutting temperature was measured with artificial thermocouples and question of investigation of metal machinability from aspect of cutting temperature was analyzed. For investigation of material machinability during turning artificial thermocouple was placed just below the cutting top of insert, and for drilling thermocouples were placed through screw holes on the face surface. In this way was obtained simple, reliable, economic and accurate method for investigation of cutting machinability.

  9. Measurement of spin amplitudes of neutron scattering on 159Tb by neutron diffraction method in antiferromagnets at ultra-low temperature

    International Nuclear Information System (INIS)

    Akopyan, G.G.; Alfimenkov, V.P.; Lason', L.; Ovchinnikov, O.N.; Sharapov, Eh.I.

    1975-01-01

    The measurement results are presented of the spin-spin amplitude for 159 Tb performed by the method of neutron diffraction by the flight time in the IBR-30 pulse reactor with samples cooled down to ultralow temperatures in a cryostat with 3 He dissolved in 4 He. The scattering angle has been fixed, neutrons incident upon the sample had a continuous energy spectrum, and the wave length satisfying the Bragg-Wolf condition has been determined by the neutron flight time required for given distance. Small-sized helium counters served as detectors. Preliminary measurements have been done with terbium dioxide and cobalt oxide whose intraatomic field values are not high. For main measurements TbAg intermetal has been chosen in which the magnetic field on a terbium nuclei is much greater. The difference of spin amplitudes of the thermal neutron scattering by terbium nuclei appeared to be equal to αsub(+-)αsub(-)=(0.35+-0.14) fermi. The spin-noncoherent scattering cross section has been obtained to be sigmasub(inc)=5+-4 mbn. The experiments have shown that the method is highly sensitive to a temperature

  10. Empirical Temperature Measurement in Protoplanetary Disks

    Science.gov (United States)

    Weaver, Erik; Isella, Andrea; Boehler, Yann

    2018-02-01

    The accurate measurement of temperature in protoplanetary disks is critical to understanding many key features of disk evolution and planet formation, from disk chemistry and dynamics, to planetesimal formation. This paper explores the techniques available to determine temperatures from observations of single, optically thick molecular emission lines. Specific attention is given to issues such as the inclusion of optically thin emission, problems resulting from continuum subtraction, and complications of real observations. Effort is also made to detail the exact nature and morphology of the region emitting a given line. To properly study and quantify these effects, this paper considers a range of disk models, from simple pedagogical models to very detailed models including full radiative transfer. Finally, we show how the use of the wrong methods can lead to potentially severe misinterpretations of data, leading to incorrect measurements of disk temperature profiles. We show that the best way to estimate the temperature of emitting gas is to analyze the line peak emission map without subtracting continuum emission. Continuum subtraction, which is commonly applied to observations of line emission, systematically leads to underestimation of the gas temperature. We further show that once observational effects such as beam dilution and noise are accounted for, the line brightness temperature derived from the peak emission is reliably within 10%–15% of the physical temperature of the emitting region, assuming optically thick emission. The methodology described in this paper will be applied in future works to constrain the temperature, and related physical quantities, in protoplanetary disks observed with ALMA.

  11. Measurement of very rapidly variable temperatures

    International Nuclear Information System (INIS)

    Elberg, S.; Mathonnet, P.

    1974-01-01

    Bibliographical research and visits to laboratories were undertaken in order to survey the different techniques used to measure rapidly variable temperatures, specifying the limits in maximum temperature and variation rate (time constant). On the basis of the bibliographical study these techniques were classified in three categories according to the physical meaning of their response time. Extension of the bibliographical research to methods using fast temperature variation measurement techniques and visits to research and industrial laboratories gave in an idea of the problems raised by the application of these methods. The use of these techniques in fields other than those for which they were developed can sometimes be awkward in the case of thermometric probe devices where the time constant cannot generally be specified [fr

  12. Time-Dependent Temperature Measurements in Post-Detonation Combustion: Current State-of-the-Art Methods and Emerging Technologies

    Science.gov (United States)

    2016-03-01

    becquerel (Bq)] roentgen (R) [air exposure] 2.579 760 × 10 –4 coulomb per kilogram (C kg –1 ) rad [ absorbed dose] 1 × 10 –2 joule per kilogram...in the IR. Many of the same species mention above (CO, CO2 and H2O) are present in small concentrations in air and absorb the light emitted by the...work in this area has been performed with FTIR spectrometers which have measurements rates typically 50 Hz or less and cost >$100k. Faster data

  13. Characteristics of Evapotranspiration of Urban Lawns in a Sub-Tropical Megacity and Its Measurement by the ‘Three Temperature Model + Infrared Remote Sensing’ Method

    Directory of Open Access Journals (Sweden)

    Guoyu Qiu

    2017-05-01

    Full Text Available Evapotranspiration (ET is one of the most important factors in urban water and energy regimes. Because of the extremely high spatial heterogeneity of urban area, accurately measuring ET using conventional methods remains a challenge due to their fetch requirements and low spatial resolution. The goals of this study were to investigate the characteristics of urban ET and its main influencing factors and subsequently to improve a fetch-free, high spatial resolution method for urban ET estimation. The Bowen ratio and the ‘three-temperature model (3T model + infrared remote sensing (RS’ methods were used for these purposes. The results of this study are listed in the following lines. (1 Urban ET is mainly affected by solar radiation and the effects of air humidity, wind velocity, and air temperature are very weak; (2 The average daily, monthly, and annual ETs of the urban lawn are 2.70, 60–100, and 990 mm, respectively, which are obvious compared with other landscapes; (3 The ratio of ET to precipitation is 0.65 in the wet season and 2.6 in the dry season, indicating that most of the precipitation is evaporated; (4 The fetch-free approach of ‘3T model + infrared RS’ is verified to be an accurate method for measuring urban ET and it agrees well with the Bowen ratio method (R2 is over 0.93 and the root mean square error is less than 0.04 mm h−1; (5 The spatial heterogeneity of urban ET can also be accurately estimated by the proposed approach. These results are helpful for improving the accuracy of ET estimation in urban areas and are useful for urban water and environmental planning and management.

  14. MIVOC method with temperature controla)

    Science.gov (United States)

    Takasugi, W.; Wakaisami, M.; Sasaki, N.; Sakuma, T.; Yamamoto, M.; Kitagawa, A.; Muramatsu, M.

    2010-02-01

    The Heavy Ion Medical Accelerator in Chiba at the National Institute of Radiological Sciences has been used for cancer therapy, physics, and biology experiments since 1994. Its ion sources produce carbon ion for cancer therapy. They also produce various ions (H+-Xe21+) for physics and biology experiments. Most ion species are produced from gases by an 18 GHz electron cyclotron resonance ion source. However, some of ion species is difficult to produce from stable and secure gases. Such ion species are produced by the sputtering method. However, it is necessary to reduce material consumption rate as much as possible in the case of rare and expensive stable isotopes. We have selected "metal ions from volatile compounds method" as a means to solve this problem. We tested a variety of compounds. Since each compound has a suitable temperature to obtain the optimum vapor pressure, we have developed an accurate temperature control system. We have produced ions such as F58e9+, Co9+, Mg5+, Ti10+, Si5+, and Ge12+ with the temperature control.

  15. Temperature measurement of accelerator cell solenoid loop

    International Nuclear Information System (INIS)

    Mu Fan; Dong Pan; Dai Zhiyong

    2010-01-01

    This paper presents the research on temperature measurement of solenoid loop. The measuring temperature fiber is layered in solenoid loop for the accelerator cell. When the solenoid loop is supplied with high current form a constant current source, its temperature increases rapidly. The temperature fiber can measure the temperature of the solenoid loop and get temperature measurement rule. Experiment and simulation show temperature of interior solenoid loop the highest and it decreases from the interior to the exterior of solenoid loop. To control temperature of solenoid loop under 60 degree C, simulation displays load interval of constant current source with 80 A current should be at least is 17.5 minutes. (authors)

  16. Laser Pyrometer For Spot Temperature Measurements

    Science.gov (United States)

    Elleman, D. D.; Allen, J. L.; Lee, M. C.

    1988-01-01

    Laser pyrometer makes temperature map by scanning measuring spot across target. Scanning laser pyrometer passively measures radiation emitted by scanned spot on target and calibrated by similar passive measurement on blackbody of known temperature. Laser beam turned on for active measurements of reflectances of target spot and reflectance standard. From measurements, temperature of target spot inferred. Pyrometer useful for non-contact measurement of temperature distributions in processing of materials.

  17. Application of digital holography in temperature distribution measurement

    Science.gov (United States)

    Wang, Guangjun; Li, Yan; Wang, Dayong; Zhao, Jie

    2010-11-01

    A reflection heat source including a radiator as well as an aluminum plate is designed, and the temperature field of the aluminum plate is used as the tested object. The reflection lensless Fourier transform (LFT) digital holography is performed to measure the temperature field distribution. For the comparison, the temperature measurement system within the radiator is used to measure the temperature distributions. The results obtained by these two methods are in good agreement, which demonstrates that the digital holography method is valid for the measurement of the temperature distribution.

  18. High temperature measurement of water vapor absorption

    Science.gov (United States)

    Keefer, Dennis; Lewis, J. W. L.; Eskridge, Richard

    1985-01-01

    An investigation was undertaken to measure the absorption coefficient, at a wavelength of 10.6 microns, for mixtures of water vapor and a diluent gas at high temperature and pressure. The experimental concept was to create the desired conditions of temperature and pressure in a laser absorption wave, similar to that which would be created in a laser propulsion system. A simplified numerical model was developed to predict the characteristics of the absorption wave and to estimate the laser intensity threshold for initiation. A non-intrusive method for temperature measurement utilizing optical laser-beam deflection (OLD) and optical spark breakdown produced by an excimer laser, was thoroughly investigated and found suitable for the non-equilibrium conditions expected in the wave. Experiments were performed to verify the temperature measurement technique, to screen possible materials for surface initiation of the laser absorption wave and to attempt to initiate an absorption wave using the 1.5 kW carbon dioxide laser. The OLD technique was proven for air and for argon, but spark breakdown could not be produced in helium. It was not possible to initiate a laser absorption wave in mixtures of water and helium or water and argon using the 1.5 kW laser, a result which was consistent with the model prediction.

  19. Automated Methods Of Corrosion Measurements

    DEFF Research Database (Denmark)

    Bech-Nielsen, Gregers; Andersen, Jens Enevold Thaulov; Reeve, John Ch

    1997-01-01

    The chapter describes the following automated measurements: Corrosion Measurements by Titration, Imaging Corrosion by Scanning Probe Microscopy, Critical Pitting Temperature and Application of the Electrochemical Hydrogen Permeation Cell.......The chapter describes the following automated measurements: Corrosion Measurements by Titration, Imaging Corrosion by Scanning Probe Microscopy, Critical Pitting Temperature and Application of the Electrochemical Hydrogen Permeation Cell....

  20. Uncertainty evaluation in transition temperature measurements

    Energy Technology Data Exchange (ETDEWEB)

    Brillaud, C. [Electricite de France, Avoine (France); Augendre, H. [Electricite de France, Clamart (France); Bethmont, M. [Electricite de France, Ecuelles (France)

    1996-12-31

    The pressure vessel surveillance program is mainly based on the transition temperature change assessment, a change which is induced by neutron irradiation. Uncertainties in Charpy test measurements are well known; however, the authors are less familiar with uncertainties due to general procedures governing experiments, which can be significant and therefore must be taken into account. In fact, procedures specify neither the number of specimens needed to obtain a transition curve, nor the choice of test temperatures, nor the fitting method for the transition curve. A study has been conducted to determine the influence of the experimental procedure on the accuracy of transition temperature determination, and the initial results are presented in this paper. Two EDF laboratories performed Charpy tests on the surveillance program reference metal, using 8, 16, 24, 32 and 64 specimens to evaluate how the number of specimens affects the transition temperature. The influence of the scatter of mechanical properties has also been studied at two levels of irradiation. The authors have evaluated the effect of different sampling strategies and investigated a new fitting method, which is based on a simultaneous fitting of all curves with common constraints on parameters.

  1. Temperature measurement by neutron resonance radiography

    International Nuclear Information System (INIS)

    Mayers, J.; Baciocco, G.; Hannon, A.C.

    1988-03-01

    We present a new data analysis technique for obtaining temperatures from neutron resonance radiography measurements. The technique is applied to measurements on tantalum and rhenium foils, a high temperature engineering alloy and a model catalytic system. Temperatures have been obtained to an accuracy of better than ± 1 0 C in the temperature range 21 0 C to 600 0 C. The results are in good agreement with theoretical simulations. (author)

  2. Automatic temperature control method of shipping can

    International Nuclear Information System (INIS)

    Nishikawa, Kaoru.

    1992-01-01

    The present invention provides a method of rapidly and accurately controlling the temperature of a shipping can, which is used upon shipping inspection for a nuclear fuel assembly. That is, a measured temperature value of the shipping can is converted to a gas pressure setting value in a jacket of the shipping can by conducting a predetermined logic calculation by using a fuzzy logic. A gas pressure control section compares the pressure setting value of a fuzzy estimation section and the measured value of the gas pressure in the jacket of the shipping can, and conducts air supply or exhaustion of the jacket gas so as to adjust the measured value with the setting value. These fuzzy estimation section and gas pressure control section control the gas pressure in the jacket of the shipping can to control the water level in the jacket. As a result, the temperature of the shipping can is controlled. With such procedures, since the water level in the jacket can be controlled directly and finely, temperature of the shipping can is automatically controlled rapidly and accurately compared with a conventional case. (I.S.)

  3. Influence of Sensor Ingestion Timing on Consistency of Temperature Measures

    National Research Council Canada - National Science Library

    Goodman, Daniel A; Kenefick, Robert W; Cadarette, Bruce S; Cheuvront, Samuel N

    2009-01-01

    ... (ITS) to measure core body temperature have been demonstrated. However, the effect of elapsed time between ITS ingestion and Tint measurement has not been thoroughly studied. Methods: Eight volunteers...

  4. Temperature measurements in cavitation bubbles

    Science.gov (United States)

    Coutier-Delgosha, Olivier

    2016-11-01

    Cavitation is usually a nearly isothermal process in the liquid phase, but in some specific flow conditions like hot water or cryogenic fluids, significant temperature variations are detected. In addition, a large temperature increase happens inside the cavitation bubbles at the very end of their collapse, due to the fast compression of the gas at the bubble core, which is almost adiabatic. This process is of primary interest in various biomedical and pharmaceutical applications, where the mechanisms of bubble collapse plays a major role. To investigate the amplitude and the spatial distribution of these temperature variations inside and outside the cavitation bubbles, a system based on cold wires has been developed. They have been tested in a configuration of a single bubble obtained by submitting a small air bubble to a large amplitude pressure wave. Some promising results have been obtained after the initial validation tests. This work is funded by the Office of Naval Research Global under Grant N62909-16-1-2116, Dr. Salahuddin Ahmed & Ki-Han Kim program managers.

  5. Quantitative shearography in axisymmetric gas temperature measurements

    Science.gov (United States)

    VanDerWege, Brad A.; O'Brien, Christopher J.; Hochgreb, Simone

    1999-06-01

    This paper describes the use of shearing interferometry (shearography) for the quantitative measurement of gas temperatures in axisymmetric systems in which vibration and shock are substantial, and measurement time is limited. The setup and principle of operation of the interferometer are described, as well as Fourier-transform-based fringe pattern analysis, Abel transform, and sensitivity of the phase lead to temperature calculation. A helium jet and a Bunsen burner flame are shown as verification of the diagnostic. The accuracy of the measured temperature profile is shown to be limited by the Abel transform and is critically dependent on the reference temperature used.

  6. Minimizing noise-temperature measurement errors

    Science.gov (United States)

    Stelzried, C. T.

    1992-01-01

    An analysis of noise-temperature measurement errors of low-noise amplifiers was performed. Results of this analysis can be used to optimize measurement schemes for minimum errors. For the cases evaluated, the effective noise temperature (Te) of a Ka-band maser can be measured most accurately by switching between an ambient and a 2-K cooled load without an isolation attenuator. A measurement accuracy of 0.3 K was obtained for this example.

  7. High-temperature thermocouples and related methods

    Science.gov (United States)

    Rempe, Joy L [Idaho Falls, ID; Knudson, Darrell L [Firth, ID; Condie, Keith G [Idaho Falls, ID; Wilkins, S Curt [Idaho Falls, ID

    2011-01-18

    A high-temperature thermocouple and methods for fabricating a thermocouple capable of long-term operation in high-temperature, hostile environments without significant signal degradation or shortened thermocouple lifetime due to heat induced brittleness.

  8. Temperature Sensitive Particle for Velocity and Temperature Measurement.

    Science.gov (United States)

    Someya, Satoshi; Okamoto, Koji; Iida, Masao

    2007-11-01

    Phosphorescence and fluorescence are often applied to measure the temperature and the concentration of oxygen. The intensity and the lifetime of phosphor depend on the temperature and the oxygen concentration, due to the quenching effect of the phosphor. The present study clarified the effects of temperature on the lifetime of phosphorescence of Porphyrins, Ru(bpy)3^2+ and the europium complex. The phosphorescence lifetime of oil solution / water solution / painted wall were measured with changing temperature and oxygen concentration. In addition, the optical property of the small particles incorporated with the europium complex was investigated in the oil/water. The lifetime was strongly affected by temperature. Then, the temperature sensitive particle (TSParticle) with metal complex was applied to measure temperature in Silicone oil (10cSt) two-dimensionally. Present study is the result of ?High speed three-dimensional direct measurement technology development for the evaluation of heat flux and flow of liquid metal? entrusted to the University of Tokyo by the Ministry of Education, Culture, Sports, Science and Technology of Japan(MEXT).

  9. Nanosecond-resolved temperature measurements using magnetic nanoparticles

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Wenbiao; Zhang, Pu [School of Automation, Huazhong University of Science and Technology, Wuhan 430074 (China); Liu, Wenzhong, E-mail: lwz7410@hust.edu.cn [School of Automation, Huazhong University of Science and Technology, Wuhan 430074 (China); Key Laboratory of Image Processing and Intelligent Control, Huazhong University of Science and Technology, Wuhan 430074 (China)

    2016-05-15

    Instantaneous and noninvasive temperature measurements are important when laser thermotherapy or welding is performed. A noninvasive nanosecond-resolved magnetic nanoparticle (MNP) temperature measurement system is described in which a transient change in temperature causes an instantaneous change in the magnetic susceptibilities of the MNPs. These transient changes in the magnetic susceptibilities are rapidly recorded using a wideband magnetic measurement system with an upper frequency limit of 0.5 GHz. The Langevin function (the thermodynamic model characterizing the MNP magnetization process) is used to obtain the temperature information. Experiments showed that the MNP DC magnetization temperature-measurement system can detect a 14.4 ns laser pulse at least. This method of measuring temperature is likely to be useful for acquiring the internal temperatures of materials irradiated with lasers, as well as in other areas of research.

  10. Nanosecond-resolved temperature measurements using magnetic nanoparticles

    Science.gov (United States)

    Xu, Wenbiao; Liu, Wenzhong; Zhang, Pu

    2016-05-01

    Instantaneous and noninvasive temperature measurements are important when laser thermotherapy or welding is performed. A noninvasive nanosecond-resolved magnetic nanoparticle (MNP) temperature measurement system is described in which a transient change in temperature causes an instantaneous change in the magnetic susceptibilities of the MNPs. These transient changes in the magnetic susceptibilities are rapidly recorded using a wideband magnetic measurement system with an upper frequency limit of 0.5 GHz. The Langevin function (the thermodynamic model characterizing the MNP magnetization process) is used to obtain the temperature information. Experiments showed that the MNP DC magnetization temperature-measurement system can detect a 14.4 ns laser pulse at least. This method of measuring temperature is likely to be useful for acquiring the internal temperatures of materials irradiated with lasers, as well as in other areas of research.

  11. Heat transfer measurements with TOIRT method

    Science.gov (United States)

    Solnař, S.; Petera, K.; Dostál, M.; Jirout, T.

    Temperature Oscillation Infra-Red Thermography (TOIRT) method was used to measure heat transfer coefficients between a at surface and a confined impinging jet generated by an impeller in a difusor and baffled vessel. The TOIRT method is based on measuring a phase-lag between the oscillating heat flux applied to the heat transfer surface and the surface temperature response using a contactless infra-red camera. The phase lag is in a direct relationship with the heat transfer coefficient.

  12. Estimation of complete temperature fields from measured temperatures

    International Nuclear Information System (INIS)

    Clegg, S.T.; Roemer, R.B.

    1984-01-01

    In hyperthermia treatments, it is desirable to be able to predict complete tissue temperature fields from sampled temperatures taken at a few locations. This is a difficult problem in hyperthermia treatments since the tissue blood perfusion is unknown. An initial attempt to do this automatically using unconstrained optimization techniques to minimize the differences between steady state temperatures measured during a treatment and temperatures (at the same locations) predicted from treatment simulations has been previously reported. A second technique using transient temperatures following a step decrease in power has been developed. This technique, which appears to be able to better predict complete temperature fields is presented and both it and the steady state technique are applied to data from both simulated and experimental hyperthermia treatments. The results of applying the two techniques are compared for one-dimensional situations. One particularly important problem which the transient technique can solve (and the steady state technique does not seem to be able to do as well) is that of predicting the complete temperature field in situations where the true maximum and/or minimum temperatures present are not measured by the available instrumentation

  13. [Temperature Measurement with Bluetooth under Android Platform].

    Science.gov (United States)

    Wang, Shuai; Shen, Hao; Luo, Changze

    2015-03-01

    To realize the real-time transmission of temperature data and display using the platform of intelligent mobile phone and bluetooth. Application of Arduino Uno R3 in temperature data acquisition of digital temperature sensor DS18B20 acquisition, through the HC-05 bluetooth transmits the data to the intelligent smart phone Android system, realizes transmission of temperature data. Using Java language to write applications program under Android development environment, can achieve real-time temperature data display, storage and drawing temperature fluctuations drawn graphics. Temperature sensor is experimentally tested to meet the body temperature measurement precision and accuracy. This paper can provide a reference for other smart phone mobile medical product development.

  14. How is it possible to measure a nuclear temperature

    International Nuclear Information System (INIS)

    Tamain, B.

    1989-01-01

    Several methods for the measurement of nuclear temperatures are summarized. The concepts of hot nuclei and temperature are defined. The nuclear equation of state is presented. The statistical theory of hot nuclei decay properties is analyzed. The obtention of the excitation energy from the recoil velocity measurement is considered in the case of complete and incomplete fusion. The measurements of temperature and excitation energy from the properties of decay products are reviewed. The study shows that no measurement method is perfect. Moreover, it is necessary to select events for which the degree of dissipation of the incident energy is estimated

  15. Dynamic Inertia Measurement Method

    Data.gov (United States)

    National Aeronautics and Space Administration — Critically important inertia measurements are complex and expensive to obtain due to the extensive fixturing and custom instrumentation of conventional techniques....

  16. MPPT Technique Based on Current and Temperature Measurements

    OpenAIRE

    Vicente, Eduardo Moreira; Moreno, Robson Luiz; Ribeiro, Enio Roberto

    2015-01-01

    This paper presents a new maximum power point tracking (MPPT) method based on the measurement of temperature and short-circuit current, in a simple and efficient approach. These measurements, which can precisely define the maximum power point (MPP), have not been used together in other existing techniques. The temperature is measured with a low cost sensor and the solar irradiance is estimated through the relationship of the measured short-circuit current and its reference. Fast tracking spee...

  17. Solar cell junction temperature measurement of PV module

    KAUST Repository

    Huang, B.J.

    2011-02-01

    The present study develops a simple non-destructive method to measure the solar cell junction temperature of PV module. The PV module was put in the environmental chamber with precise temperature control to keep the solar PV module as well as the cell junction in thermal equilibrium with the chamber. The open-circuit voltage of PV module Voc is then measured using a short pulse of solar irradiation provided by a solar simulator. Repeating the measurements at different environment temperature (40-80°C) and solar irradiation S (200-1000W/m2), the correlation between the open-circuit voltage Voc, the junction temperature Tj, and solar irradiation S is derived.The fundamental correlation of the PV module is utilized for on-site monitoring of solar cell junction temperature using the measured Voc and S at a short time instant with open circuit. The junction temperature Tj is then determined using the measured S and Voc through the fundamental correlation. The outdoor test results show that the junction temperature measured using the present method, Tjo, is more accurate. The maximum error using the average surface temperature Tave as the junction temperature is 4.8 °C underestimation; while the maximum error using the present method is 1.3 °C underestimation. © 2010 Elsevier Ltd.

  18. Dynamic temperature measurements with embedded optical sensors.

    Energy Technology Data Exchange (ETDEWEB)

    Dolan, Daniel H.,; Seagle, Christopher T; Ao, Tommy

    2013-10-01

    This report summarizes LDRD project number 151365, \\Dynamic Temperature Measurements with Embedded Optical Sensors". The purpose of this project was to develop an optical sensor capable of detecting modest temperature states (<1000 K) with nanosecond time resolution, a recurring diagnostic need in dynamic compression experiments at the Sandia Z machine. Gold sensors were selected because the visible re ectance spectrum of gold varies strongly with temperature. A variety of static and dynamic measurements were performed to assess re ectance changes at di erent temperatures and pressures. Using a minimal optical model for gold, a plausible connection between static calibrations and dynamic measurements was found. With re nements to the model and diagnostic upgrades, embedded gold sensors seem capable of detecting minor (<50 K) temperature changes under dynamic compression.

  19. Electron Density and Temperature Measurements, and Abundance ...

    Indian Academy of Sciences (India)

    2016-01-27

    Jan 27, 2016 ... Using spectra obtained from the SUMER (Solar Ultraviolet Measurements of Emitted Radiation) spectrograph on the spacecraft SOHO (Solar and Heliospheric Observatory), we investigate the height dependence of electron density, temperature and abundance anomalies in the solar atmosphere.

  20. General temperature field measurement by digital holography

    Czech Academy of Sciences Publication Activity Database

    Doleček, Roman; Psota, Pavel; Lédl, Vít; Vít, Tomáš; Václavík, Jan; Kopecký, V.

    2013-01-01

    Roč. 52, č. 1 (2013), A319-A325 ISSN 1559-128X Institutional support: RVO:61389021 Keywords : digital holography * temperature field measurement * tomography Subject RIV: BH - Optics, Masers, Lasers Impact factor: 1.649, year: 2013

  1. Investigation into Methods for Predicting Connection Temperatures

    Directory of Open Access Journals (Sweden)

    K. Anderson

    2009-01-01

    Full Text Available The mechanical response of connections in fire is largely based on material strength degradation and the interactions between the various components of the connection. In order to predict connection performance in fire, temperature profiles must initially be established in order to evaluate the material strength degradation over time. This paper examines two current methods for predicting connection temperatures: The percentage method, where connection temperatures are calculated as a percentage of the adjacent beam lower-flange, mid-span temperatures; and the lumped capacitance method, based on the lumped mass of the connection. Results from the percentage method do not correlate well with experimental results, whereas the lumped capacitance method shows much better agreement with average connection temperatures. A 3D finite element heat transfer model was also created in Abaqus, and showed good correlation with experimental results. 

  2. Wideband filter radiometers for blackbody temperature measurements

    Science.gov (United States)

    Boivin, L. P.; Bamber, C.; Gaertner, A. A.; Gerson, R. K.; Woods, D. J.; Woolliams, E. R.

    2010-10-01

    The use of high-temperature blackbody (HTBB) radiators to realize primary spectral irradiance scales requires that the operating temperature of the HTBB be accurately determined. We have developed five filter radiometers (FRs) to measure the temperature of the National Research Council of Canada's HTBB. The FRs are designed to minimize sensitivity to ambient temperature fluctuations. They incorporate air-spaced colored glass filters and a Si photodiode detector that are housed in a cell whose temperature is controlled to ±0.1°C by means of annular thermoelectric elements at the front and rear of the cell. These wideband filter radiometers operate in four different wavelength bands. The spectral responsivity measurements were performed in an underfill geometry for a power-mode calibration that is traceable to NRC's cryogenic radiometer. The spectral temperature sensitivity of each of these FRs has been measured. The apertures for these FRs were cold-formed by swaging machine-cut apertures onto precision dowel pins. A description of the filter radiometer design, fabrication and testing, together with a detailed uncertainty analysis, is presented. We derive the equations that relate the spectral irradiance measured by the FRs to the spectral radiance and temperature of the HTBB, and deal specifically with the change of index of refraction over the path of the radiation from the interior of the HTBB to the FRs. We believe these equations are more accurate than recently published derivations. Our measurements of the operating temperature of our HTBB working at temperatures near 2500 K, 2700 K and 2900 K, together with measurements using a pyrometer, show agreement between the five filter radiometers and with the pyrometer to within the estimated uncertainties.

  3. Online junction temperature measurement using peak gate current

    DEFF Research Database (Denmark)

    Baker, Nick; Munk-Nielsen, Stig; Iannuzzo, Francesco

    2015-01-01

    A new method for junction temperature measurement of MOS-gated power semiconductor switches is presented. The measurement method involves detecting the peak voltage over the external gate resistor of an IGBT or MOSFET during turn-on. This voltage is directly proportional to the peak gate current...

  4. Improvement of the Method for Reconstructing the Temperature and Salinity Three-Dimensional Fields of the Black Sea Based on Insufficient Measurements and Altimetry

    Directory of Open Access Journals (Sweden)

    V.V. Knysh

    2016-12-01

    Full Text Available The article represents the results of two special numerical experiments aimed at improving the previously proposed procedure of reconstructing salinity and temperature three-dimensional fields based on the altimetry data and the insufficient measurements performed at the stations and the Argo buoys in 2012. In the Experiment 1, the monthly average coefficients of sea level linear dependence and depths where the salinity values of the “zero” gradation altimetry level profile lie within the salinity profiles of positive and negative gradations are applied. The procedure for calculating the daily average coefficients of the depth linear trends is realized in the Experiment 2. It is shown that the thermohaline fields reconstructed in the Experiment 2 for the deepwater area are more accurate; on the horizons of the 100–500 m layer their values range smoothly from one day to another. The Black Sea hydrophysical fields are reconstructed by assimilation in the model of three-dimensional thermohaline parameters in the reanalysis for 2012. It is revealed that, as compared to the observations on the overwhelming majority of horizons in the 0–500 m layer (the Experiment 2, the standard root-mean-square deviations of temperature and salinity are lower than those in Experiment 1. The root of the measured salinity field dispersion exceeds the standard deviations on all the horizons within 0–500 m, inclusive. Application of the daily average coefficients of the linear trends for reconstructing three-dimensional fields of temperature and salinity, and their subsequent assimilation in the model is preferable. It is revealed that the model of the upper 0–100 m layer thermodynamics requires improvement.

  5. Assessment of body temperature measurement options.

    Science.gov (United States)

    Sund-Levander, Märtha; Grodzinsky, Ewa

    Assessment of body temperature is important for decisions in nursing care, medical diagnosis, treatment and the need of laboratory tests. The definition of normal body temperature as 37°C was established in the middle of the 19th century. Since then the technical design and the accuracy of thermometers has been much improved. Knowledge of physical influence on the individual body temperature, such as thermoregulation and hormones, are still not taken into consideration in body temperature assessment. It is time for a change; the unadjusted mode should be used, without adjusting to another site and the same site of measurement should be used as far as possible. Peripheral sites, such as the axillary and the forehead site, are not recommended as an assessment of core body temperature in adults. Frail elderly individuals might have a low normal body temperature and therefore be at risk of being assessed as non-febrile. As the ear site is close to the hypothalamus and quickly responds to changes in the set point temperature, it is a preferable and recommendable site for measurement of body temperature.

  6. Spectroscopic analysis applied to temperature measurement in plasmas

    International Nuclear Information System (INIS)

    Fieffe-Prevost, P.

    1978-01-01

    The plasma temperature is defined only if the plasma is in a state near thermodynamic equilibrium. This plasma state is analysed in detail and spectroscopic methods for measuring the temperature are discussed. As an application the hydrogen arc of the National Institute of Metrology of the Conservatoire National des Arts et Metiers (Paris) is briefly described [fr

  7. Survey of Temperature Measurement Techniques For Studying Underwater Shock Waves

    Science.gov (United States)

    Danehy, Paul M.; Alderfer, David W.

    2004-01-01

    Several optical methods for measuring temperature near underwater shock waves are reviewed and compared. The relative merits of the different techniques are compared, considering accuracy, precision, ease of use, applicable temperature range, maturity, spatial resolution, and whether or not special additives are required.

  8. Matter and Methods at Low Temperatures

    CERN Document Server

    Pobell, F

    2007-01-01

    Matter and Methods at Low Temperatures contains a wealth of information essential for successful experiments at low temperatures, which makes it suitable as a reference and textbook. The first chapters describe the low-temperature properties of liquid and solid matter, including liquid helium. The major part of the book is devoted to refrigeration techniques and the physics on which they rely, the definition of temperature, thermometry, and a variety of design and construction techniques. The lively style and practical basis of this text make it easy to read and particularly useful to anyone beginning research in low-temperature physics. Low-temperature scientists will find it of great value due to its extensive compilation of materials data and relevant new results on refrigeration, thermometry, and materials properties. Problems are included as well. Furthermore, this third edition also describes newly developed low-temperature experimentation techniques and new materials properties; it also contains many a...

  9. Temperature lags of luminescence measurements in a commercial luminescence reader

    Energy Technology Data Exchange (ETDEWEB)

    Kitis, George [Aristotle University of Thessaloniki, Nuclear Physics Laboratory, 54124 Thessaloniki (Greece); Kiyak, Nafiye G. [ISIK University, Faculty of Science and Arts, Physics Department, Sile, 34980 Istanbul (Turkey); Polymeris, George S., E-mail: gspolymeris@ankara.edu.tr [Ankara University, Institute of Nuclear Sciences, Beşevler, 06100 Ankara (Turkey)

    2015-09-15

    The temperature recorded in thermoluminescence and optically stimulated luminescence equipments is not the temperature of the sample but that of the heating element on which the thermocouple is attached. Depending upon the rate of heating, a temperature difference appears between the samples and the heating element, termed as temperature lag, which could have serious effects on the curve shapes and trapping parameters. In the present work the temperature lag effect is studied in a newly developed luminescence equipment measuring both thermoluminescence and optically stimulated luminescence. It is found that the temperature lag could be large for heating rates above 2 K/s and it is strongly dependent upon the sample holder. A simple approximation method is proposed in order to both predict as well as correct for temperature lag effects in luminescence measurements.

  10. Turbine gas temperature measurement and control system

    Science.gov (United States)

    Webb, W. L.

    1973-01-01

    A fluidic Turbine Inlet Gas Temperature (TIGIT) Measurement and Control System was developed for use on a Pratt and Whitney Aircraft J58 engine. Based on engine operating requirements, criteria for high temperature materials selection, system design, and system performance were established. To minimize development and operational risk, the TIGT control system was designed to interface with an existing Exhaust Gas Temperature (EGT) Trim System and thereby modulate steady-state fuel flow to maintain a desired TIGT level. Extensive component and system testing was conducted including heated (2300F) vibration tests for the fluidic sensor and gas sampling probe, temperature and vibration tests on the system electronics, burner rig testing of the TIGT measurement system, and in excess of 100 hours of system testing on a J58 engine. (Modified author abstract)

  11. High-temperature archeointensity measurements from Mesopotamia

    Science.gov (United States)

    Gallet, Yves; Le Goff, Maxime

    2006-01-01

    We present new archeointensity results obtained from 127 potsherds and baked brick fragments dated from the last four millennia BC which were collected from different Syrian archeological excavations. High temperature magnetization measurements were carried out using a laboratory-built triaxial vibrating sample magnetometer (Triaxe), and ancient field intensity determinations were derived from the experimental procedure described by Le Goff and Gallet [Le Goff and Gallet. Earth Planet. Sci. Lett. 229 (2004) 31-43]. As some of the studied samples were previously analyzed using the classical Thellier and Thellier [Thellier and Thellier . Ann. Geophys. 15 (1959) 285-376] method revised by Coe [Coe. J. Geophys. Res. 72 (1967) 3247-3262], a comparison of the results is made from the two methods. The differences both at the fragment and site levels are mostly within ± 5%, which strengthens the validity of the experimental procedure developed for the Triaxe. The new data help to better constrain the geomagnetic field intensity variations in Mesopotamia during archeological times, with the probable occurrence of an archeomagnetic jerk around 2800-2600 BC.

  12. Laboratory setup for temperature and humidity measurements

    CERN Document Server

    Eimre, Kristjan

    2015-01-01

    In active particle detectors, the temperature and humidity conditions must be under constant monitoring and control, as even small deviations from the norm cause changes to detector characteristics and result in a loss of precision. To monitor the temperature and humidity, different kinds of sensors are used, which must be calibrated beforehand to ensure their accuracy. To calibrate the large number of sensors that are needed for the particle detectors and other laboratory work, a calibration system is needed. The purpose of the current work was to develop a laboratory setup for temperature and humidity sensor measurements and calibration.

  13. Temperature measurements of shock-compressed deuterium

    International Nuclear Information System (INIS)

    Holmes, N.C.; Ross, M.; Nellis, W.J.

    1994-11-01

    The authors measured the temperatures of single and double-shocked D 2 and H 2 up to 85 GPa (0.85 Mbar) and 5,200 K. While single shock temperatures, at pressures to 23 GPa, agree well with previous models, the double shock temperatures are as much as 40% lower than predicted. This is believed to be caused by molecular dissociation, and a new model of the hydrogen EOS at extreme conditions has been developed which correctly predicts their observations. These data and model have important implications for programs which use condensed-phase hydrogen in implosion systems

  14. Study on a transient optical fiber high temperature measurement system

    Science.gov (United States)

    Cai, Lulu; Liu, Yusha; Wang, Yutian

    2009-07-01

    High temperature is one of the most important parameters in the fields of scientific research and industrial production. At present, thermocouple, thermo resistive and radiance thermometer are already technologically mature which can be adopted to measure the general temperature, but when it comes to the transient high temperature that changes pretty quickly in wretched conditions, those traditional pyrometers can not meet the requirements any more. In this paper, we designed a transient optical high temperature measurement system. First, design of the temperature measurement probe. The system took blackbody cavity sensor together with optical fiber to receive the measured signal, here, the integrated emissivity model of the blackbody cavity was established and the optimum structure parameters were confirmed. Secondly, design of the entire temperature measurement system. A contact-noncontact measurement method was applied, which is to make the blackbody cavity and the measured high-temperature source contact, the fiber probe and the blackbody cavity noncontact, as a result, the error caused by contact measurement is overcame and the precision is guaranteed at the same time. In addition, a fiber grating was introduced as the wavelength filter device which can realize the dynamic filter of narrow-band signals and reduce the impact of background light. Thirdly, signal processing. In this part, we applied labVIEW software and wavelet analysis method. All of the signal acquisition and processing were realized in the labVIEW environment. Through calling matlab in labVIEW, the signals from optical fiber detector were wavelet denoised and decomposed, thus the temperature information was extracted, and the temperature value was obtained. On basis of wavelet transformation, the paper adopted the 4dB wavelet with horizontal scale of 5 to realize the feature extraction and noise removal, parts of the signals before and after the wavelet noise removal were given and analyzed

  15. Inverse analysis of inner surface temperature history from outer surface temperature measurement of a pipe

    International Nuclear Information System (INIS)

    Kubo, S; Ioka, S; Onchi, S; Matsumoto, Y

    2010-01-01

    When slug flow runs through a pipe, nonuniform and time-varying thermal stresses develop and there is a possibility that thermal fatigue occurs. Therefore it is necessary to know the temperature distributions and the stress distributions in the pipe for the integrity assessment of the pipe. It is, however, difficult to measure the inner surface temperature directly. Therefore establishment of the estimation method of the temperature history on inner surface of pipe is needed. As a basic study on the estimation method of the temperature history on the inner surface of a pipe with slug flow, this paper presents an estimation method of the temperature on the inner surface of a plate from the temperature on the outer surface. The relationship between the temperature history on the outer surface and the inner surface is obtained analytically. Using the results of the mathematical analysis, the inverse analysis method of the inner surface temperature history estimation from the outer surface temperature history is proposed. It is found that the inner surface temperature history can be estimated from the outer surface temperature history by applying the inverse analysis method, even when it is expressed by the multiple frequency components.

  16. Surface temperature measurement of plasma facing components in tokamaks

    International Nuclear Information System (INIS)

    Amiel, Stephane

    2014-01-01

    During this PhD, the challenges on the non-intrusive surface temperature measurements of metallic plasma facing components in tokamaks are reported. Indeed, a precise material emissivity value is needed for classical infrared methods and the environment contribution has to be known particularly for low emissivities materials. Although methods have been developed to overcome these issues, they have been implemented solely for dedicated experiments. In any case, none of these methods are suitable for surface temperature measurement in tokamaks.The active pyrometry introduced in this study allows surface temperature measurements independently of reflected flux and emissivities using pulsed and modulated photothermal effect. This method has been validated in laboratory on metallic materials with reflected fluxes for pulsed and modulated modes. This experimental validation is coupled with a surface temperature variation induced by photothermal effect and temporal signal evolvement modelling in order to optimize both the heating source characteristics and the data acquisition and treatment. The experimental results have been used to determine the application range in temperature and detection wavelengths. In this context, the design of an active pyrometry system on tokamak has been completed, based on a bicolor camera for a thermography application in metallic (or low emissivity) environment.The active pyrometry method introduced in this study is a complementary technique of classical infrared methods used for thermography in tokamak environment which allows performing local and 2D surface temperature measurements independently of reflected fluxes and emissivities. (author) [fr

  17. Temperature measurements at the LMFBR core outlet

    International Nuclear Information System (INIS)

    Argous, J.P.; Berger, R.; Casejuane, R.; Fournier, C.; Girard, J.P.

    1980-04-01

    Over the last few years the temperature sensors used to measure the subassembly outlet temperature in French designed LMFBRs have been modified, basically in an effort to reduce the dispersion of the chromel-alumel thermocouple time constant, and to extend the frequency spectrum of the measurement signals by adding a steel electrode to from a stainless steel-sodium thermocouple. The result of this evolution is the temperature probe immersed in sodium which will be used in the SUPER PHENIX reactor. This paper describes the tests already completed or in progress on this probe. It also presents measurement data on the two basic probe parameters: the thermoelectric power of the stainless steel-sodium thermocouple and the time constant of the chromel-alumel thermocouple

  18. Dual neutron flux/temperature measurement sensor

    Science.gov (United States)

    Mihalczo, John T.; Simpson, Marc L.; McElhaney, Stephanie A.

    1994-01-01

    Simultaneous measurement of neutron flux and temperature is provided by a single sensor which includes a phosphor mixture having two principal constituents. The first constituent is a neutron sensitive 6LiF and the second is a rare-earth activated Y203 thermophosphor. The mixture is coated on the end of a fiber optic, while the opposite end of the fiber optic is coupled to a light detector. The detected light scintillations are quantified for neutron flux determination, and the decay is measured for temperature determination.

  19. Slot Antenna for Wireless Temperature Measurement Systems

    DEFF Research Database (Denmark)

    Acar, Öncel; Jakobsen, Kaj Bjarne

    2016-01-01

    This paper presents a novel clover-slot antenna for a surface-acoustic-wave sensor based wireless temperature measurement system. The slot is described by a parametric locus curve that has the shape of a clover. The antenna is operated at high temperatures, in rough environments, and has a 43......% fractional bandwidth at the 2.4 GHz ISM-band. The slot antenna has been optimized for excitation by a passive chip soldered onto it. Measurement results are compared with simulation results and show good agreements....

  20. Automated general temperature correction method for dielectric soil moisture sensors

    Science.gov (United States)

    Kapilaratne, R. G. C. Jeewantinie; Lu, Minjiao

    2017-08-01

    An effective temperature correction method for dielectric sensors is important to ensure the accuracy of soil water content (SWC) measurements of local to regional-scale soil moisture monitoring networks. These networks are extensively using highly temperature sensitive dielectric sensors due to their low cost, ease of use and less power consumption. Yet there is no general temperature correction method for dielectric sensors, instead sensor or site dependent correction algorithms are employed. Such methods become ineffective at soil moisture monitoring networks with different sensor setups and those that cover diverse climatic conditions and soil types. This study attempted to develop a general temperature correction method for dielectric sensors which can be commonly used regardless of the differences in sensor type, climatic conditions and soil type without rainfall data. In this work an automated general temperature correction method was developed by adopting previously developed temperature correction algorithms using time domain reflectometry (TDR) measurements to ThetaProbe ML2X, Stevens Hydra probe II and Decagon Devices EC-TM sensor measurements. The rainy day effects removal procedure from SWC data was automated by incorporating a statistical inference technique with temperature correction algorithms. The temperature correction method was evaluated using 34 stations from the International Soil Moisture Monitoring Network and another nine stations from a local soil moisture monitoring network in Mongolia. Soil moisture monitoring networks used in this study cover four major climates and six major soil types. Results indicated that the automated temperature correction algorithms developed in this study can eliminate temperature effects from dielectric sensor measurements successfully even without on-site rainfall data. Furthermore, it has been found that actual daily average of SWC has been changed due to temperature effects of dielectric sensors with a

  1. Ion temperature measurements in the Maryland Spheromak

    International Nuclear Information System (INIS)

    Gauvreau, J.L.

    1992-01-01

    Initial spectroscopic data from MS showed evidence of ion heating as deduced from the line widths of different ion species. Detailed measurements of OIV spectral emission line profiles in space and time revealed that heating takes place at early time, before spheromak formation and is occurring within the current discharge. The measured ion temperature is several times the electron temperature and cannot be explained by classical (Spitzer) resistivity. Classically, ions are expected to have lower temperatures than the electrons and therefore, lower temperatures than observed. High ion temperatures have been observed in different RFP's and Spheromaks but are usually associated with relaxation to the Taylor state and occur in the sustainment phase. During formation, the current delivered to start the discharge is not axisymmetric and as a consequence, X-points appear in the magnetic flux. A two dimensional analysis predicts that magnetic reconnection occurring at an X-point can give rise to high ion heating rates. A simple 0-dimensional calculation showed that within the first 20 μs, a conversion of mass flow kinetic energy into ion temperature could take place due to viscosity

  2. The extent of temporal smearing in surface-temperature histories derived from borehole temperature measurements

    Science.gov (United States)

    Clow, G.D.

    1992-01-01

    The ability of borehole temperature data to resolve past climatic events is investigated using Backus-Gilbert inversion methods. Two experimental approaches are considered: (1) the data consist of a single borehole temperature profile, and (2) the data consist of climatically-induced temperature transients measured within a borehole during a monitoring experiment. The sensitivity of the data's resolving power to the vertical distribution of the measurements, temperature measurement errors, the inclusion of a local meteorological record, and the duration of a monitoring experiment, are investigated. The results can be used to help interpret existing surface temperature histories derived from borehole temperature data and to optimize future experiments for the detection of climatic signals. ?? 1992.

  3. Measurement of rotational temperature at Kolhapur, India

    Directory of Open Access Journals (Sweden)

    G. K. Mukherjee

    2004-09-01

    Full Text Available Measurements of the hydroxyl rotational temperature for the (8,3 Meinel band have been reported from the observations of the ratio of the relative intensities of P1(2 and P1(4 lines of the OH(8,3 band at Kolhapur (16.8° N, 74.2° E, dip lat. 10.6° N in India during the period 1 November 2002-29 April 2003 using tilting-filter photometers. Mean values of rotational temperature have been computed for 60 nights. The monthly mean value of temperature lies in the range 194(±11-208(±18K. The mean rotational temperature obtained from all the measurements was found to be 202±15K. The results agree with other low-latitude measurements of rotational temperature using photometric airglow techniques. Quasi-periodic fluctuations with a period of about one to two hours have been prominent on many nights. Furthermore, the results show the general agreement between observations and model (MSIS-86 predictions.

  4. Measurement of rotational temperature at Kolhapur, India

    Directory of Open Access Journals (Sweden)

    G. K. Mukherjee

    2004-09-01

    Full Text Available Measurements of the hydroxyl rotational temperature for the (8,3 Meinel band have been reported from the observations of the ratio of the relative intensities of P1(2 and P1(4 lines of the OH(8,3 band at Kolhapur (16.8° N, 74.2° E, dip lat. 10.6° N in India during the period 1 November 2002-29 April 2003 using tilting-filter photometers. Mean values of rotational temperature have been computed for 60 nights. The monthly mean value of temperature lies in the range 194(±11-208(±18K. The mean rotational temperature obtained from all the measurements was found to be 202±15K. The results agree with other low-latitude measurements of rotational temperature using photometric airglow techniques. Quasi-periodic fluctuations with a period of about one to two hours have been prominent on many nights. Furthermore, the results show the general agreement between observations and model (MSIS-86 predictions.

  5. Temperature measurements in ZT-40M

    International Nuclear Information System (INIS)

    Little, E.M.; Haberstich, A.; Thomas, K.S.; Watt, R.G.

    1983-01-01

    Electron temperatures derived from Thomson scattering and ultrasoft x-ray (USXR) measurements taken before and after machine modifications are compared for ZT-40M. Modifications were made to the magnetic field windings to reduce field errors and the joints in the aluminum shell were coated with joint compound to reduce resistance and make all joints electrically uniform. These modifications resulted in increased plasma lifetime in ZT-40M from less than 10 ms to over 20 ms. Thomson scattering measurements were made with a single-point Thomson scattering apparatus. The scattered spectrum is collected by a three-grating spectrometer. The soft x rays are collected by a two-foil differential transmission system whose foil ratios may be easily varied. Before modifications the Thomson scattering and soft x-ray temperatures agreed up until 3 to 4 ms into the discharge. After this time the Thomson scattering temperature decreased slowly while the soft x-ray ''temperature'' increased rapidly. field errors resulted in Thomson scattering and USXR ''temperature'' time histories remaining fairly flat out to 10 to 11 ms, but introduced a small discrepancy (about 50 eV) in the absolute value of the temperatures. This change may be due either to the change in foil thickness used or to changes in radial temperature profiles. Profile changes may have been caused by the addition of four poloidal limiters or improvements to the magnetic field topology. After modifications the temperatures from both Thomson scattering and USXR were lower and the plasma density was higher. This is probably a result of the lower plasma-wall interaction with the new configuration

  6. Body Temperature Measurements for Metabolic Phenotyping in Mice

    Directory of Open Access Journals (Sweden)

    Carola W. Meyer

    2017-07-01

    Full Text Available Endothermic organisms rely on tightly balanced energy budgets to maintain a regulated body temperature and body mass. Metabolic phenotyping of mice, therefore, often includes the recording of body temperature. Thermometry in mice is conducted at various sites, using various devices and measurement practices, ranging from single-time probing to continuous temperature imaging. Whilst there is broad agreement that body temperature data is of value, procedural considerations of body temperature measurements in the context of metabolic phenotyping are missing. Here, we provide an overview of the various methods currently available for gathering body temperature data from mice. We explore the scope and limitations of thermometry in mice, with the hope of assisting researchers in the selection of appropriate approaches, and conditions, for comprehensive mouse phenotypic analyses.

  7. Thermoacoustic measurement of the temperature during microwave thermotherapy

    Science.gov (United States)

    Lou, Cunguang; Xing, Da; Nie, Liming

    2009-08-01

    Microwave thermotherapy (MT) has been an important treatment in oncology. The measurement of temperature during microwave thermotherapy is vital to ensure the safety of normal tissues. Thermoacoustic signals induced are temperature dependent. This phenomenon demonstrates that the thermal parameters are closely related to the generation of thermoacoustic pressure. Here we present the studies on pulsed microwave-induced thermoacoustic signals toward temperature monitoring of tissue thermotherapy. A high power pulsed microwave was used as thermoacoustic excitation source and heating source simultaneously, thermoacoustic pressure was captured by a multi-element linear transducer array. Excellent agreement was obtained between the inversion calculation results and the actual measurement temperature. The results suggest that thermoacoustic signals can be used to monitor thermotherapy temperature, and has the potential of reconstruct the temperature distribution by filter back-projection algorithm. This method has a great potential to develop into an integrated system for pulsed microwave thermotherapy and curative effect monitoring.

  8. Body Temperature Measurements for Metabolic Phenotyping in Mice

    Science.gov (United States)

    Meyer, Carola W.; Ootsuka, Youichirou; Romanovsky, Andrej A.

    2017-01-01

    Endothermic organisms rely on tightly balanced energy budgets to maintain a regulated body temperature and body mass. Metabolic phenotyping of mice, therefore, often includes the recording of body temperature. Thermometry in mice is conducted at various sites, using various devices and measurement practices, ranging from single-time probing to continuous temperature imaging. Whilst there is broad agreement that body temperature data is of value, procedural considerations of body temperature measurements in the context of metabolic phenotyping are missing. Here, we provide an overview of the various methods currently available for gathering body temperature data from mice. We explore the scope and limitations of thermometry in mice, with the hope of assisting researchers in the selection of appropriate approaches, and conditions, for comprehensive mouse phenotypic analyses. PMID:28824441

  9. Tokamak Plasmas: Electron temperature $(T_ {e}) $ measurements ...

    Indian Academy of Sciences (India)

    Home; Journals; Pramana – Journal of Physics; Volume 55; Issue 5-6. Tokamak Plasmas : Electron temperature ( T e ) measurements by Thomson scattering system. R Rajesh B Ramesh Kumar S K Varshney Manoj Kumar Chhaya Chavda Aruna Thakkar N C Patel Ajai Kumar Aditya Team. Contributed Papers Volume 55 ...

  10. Scanning optical pyrometer for measuring temperatures in hollow cathodes.

    Science.gov (United States)

    Polk, J E; Marrese-Reading, C M; Thornber, B; Dang, L; Johnson, L K; Katz, I

    2007-09-01

    Life-limiting processes in hollow cathodes are determined largely by the temperature of the electron emitter. To support cathode life assessment, a noncontact temperature measurement technique which employs a stepper motor-driven fiber optic probe was developed. The probe is driven inside the hollow cathode and collects light radiated by the hot interior surface of the emitter. Ratio pyrometry is used to determine the axial temperature profile. Thermocouples on the orifice plate provide measurements of the external temperature during cathode operation and are used to calibrate the pyrometer system in situ with a small oven enclosing the externally heated cathode. The diagnostic method and initial measurements of the temperature distribution in a hollow cathode are discussed.

  11. An anatomically realistic temperature phantom for radiofrequency heating measurements

    Science.gov (United States)

    Graedel, Nadine N.; Polimeni, Jonathan R.; Guerin, Bastien; Gagoski, Borjan; Wald, Lawrence L.

    2014-01-01

    Purpose An anthropomorphic phantom with realistic electrical properties allows for a more accurate reproduction of tissue current patterns during excitation. A temperature map can then probe the worst-case heating expected in the un-perfused case. We describe an anatomically realistic human head phantom that allows rapid 3D temperature mapping at 7 T. Methods The phantom was based on hand-labeled anatomical imaging data and consists of four compartments matching the corresponding human tissues in geometry and electrical properties. The increases in temperature resulting from radiofrequency excitation were measured with MR thermometry using a temperature sensitive contrast agent (TmDOTMA−) validated by direct fiber optic temperature measurements. Results Acquisition of 3D temperature maps of the full phantom with a temperature accuracy better than 0.1°C was achieved with an isotropic resolution of 5 mm and acquisition times of 2–4 minutes. Conclusion Our results demonstrate the feasibility of constructing anatomically realistic phantoms with complex geometries incorporating the ability to measure accurate temperature maps in the phantom. The anthropomorphic temperature phantom is expected to provide a useful tool for the evaluation of the heating effects of both conventional and parallel transmit pulses and help validate electromagnetic and temperature simulations. PMID:24549755

  12. Modern gas-based temperature and pressure measurements

    CERN Document Server

    Pavese, Franco

    2013-01-01

    This 2nd edition volume of Modern Gas-Based Temperature and Pressure Measurements follows the first publication in 1992. It collects a much larger set of information, reference data, and bibliography in temperature and pressure metrology of gaseous substances, including the physical-chemical issues related to gaseous substances. The book provides solutions to practical applications where gases are used in different thermodynamic conditions. Modern Gas-Based Temperature and Pressure Measurements, 2nd edition is the only comprehensive survey of methods for pressure measurement in gaseous media used in the medium-to-low pressure range closely connected with thermometry. It assembles current information on thermometry and manometry that involve the use of gaseous substances which are likely to be valid methods for the future. As such, it is an important resource for the researcher. This edition is updated through the very latest scientific and technical developments of gas-based temperature and pressure measurem...

  13. Automated methods of corrosion measurement

    DEFF Research Database (Denmark)

    Andersen, Jens Enevold Thaulov; Bech-Nielsen, Gregers; Reeve, John Ch

    1997-01-01

    to revise assumptions regarding the basis of the method, which sometimes leads to the discovery of as-yet unnoticed phenomena. The present selection of automated methods for corrosion measurements is not motivated simply by the fact that a certain measurement can be performed automatically. Automation...

  14. Determination of transient fluid temperature using the inverse method

    Directory of Open Access Journals (Sweden)

    Jaremkiewicz Magdalena

    2014-03-01

    Full Text Available This paper proposes an inverse method to obtain accurate measurements of the transient temperature of fluid. A method for unit step and linear rise of temperature is presented. For this purpose, the thermometer housing is modelled as a full cylindrical element (with no inner hole, divided into four control volumes. Using the control volume method, the heat balance equations can be written for each of the nodes for each of the control volumes. Thus, for a known temperature in the middle of the cylindrical element, the distribution of temperature in three nodes and heat flux at the outer surface were obtained. For a known value of the heat transfer coefficient the temperature of the fluid can be calculated using the boundary condition. Additionally, results of experimental research are presented. The research was carried out during the start-up of an experimental installation, which comprises: a steam generator unit, an installation for boiler feed water treatment, a tray-type deaerator, a blow down flashvessel for heat recovery, a steam pressure reduction station, a boiler control system and a steam header made of martensitic high alloy P91 steel. Based on temperature measurements made in the steam header using the inverse method, accurate measurements of the transient temperature of the steam were obtained. The results of the calculations are compared with the real temperature of the steam, which can be determined for a known pressure and enthalpy.

  15. Method for enrichment by dual temperature exchange

    International Nuclear Information System (INIS)

    Spevack, J.S.

    1977-01-01

    In dual temperature systems utilizing different fluid materials in liquid and gas phases separable from each other (for example H 2 O and H 2 S), the phases are contacted with each other at a relatively hot temperature. Herein combinations of method and means are provided by which the gas is conditioned by raising its temperature and humidity principally by heat derived from the cooling and dehumidification of said gas. Special provisions are made in the combinations for transferring said heat and for the conditioning of the gas with high efficiency; and for economically controlling the temperature of the condensate resulting from the dehumidification of the gas to adapt it for particular uses in the system. Method and means are provided for such liquid gas contacting systems for efficiently stripping or separating dissolved gas from the effluent liquid and returning one of the so separated materials to the system

  16. Stream temperature investigations: field and analytic methods

    Science.gov (United States)

    Bartholow, J.M.

    1989-01-01

    This document provides guidance to the user of the U.S. Fish and Wildlife Service’s Stream Network Temperature Model (SNTEMP). Planning a temperature study is discussed in terms of understanding the management objectives and ensuring that the questions will be accurately answered with the modeling approach being used. A sensitivity analysis of SNTEMP is presented to illustrate which input variables are most important in predicting stream temperatures. This information helps prioritize data collection activities, highlights the need for quality control, focuses on which parameters can be estimated rather than measured, and offers a broader perspective on management options in terms of knowing where the biggest temperature response will be felt. All of the major input variables for stream geometry, meteorology, and hydrology are discussed in detail. Each variable is defined, with guidance given on how to measure it, what kind of equipment to use, where to obtain it from another agency, and how to calculate it if the data are in a form other than that required by SNTEMP. Examples are presented for the various forms in which water temperature, discharge, and meteorological data are commonly found. Ranges of values for certain input variables that are difficult to measure of estimate are given. Particular attention is given to those variables not commonly understood by field biologists likely to be involved in a stream temperature study. Pertinent literature is cited for each variable, with emphasis on how other people have treated particular problems and on results they have found.

  17. MPPT Technique Based on Current and Temperature Measurements

    Directory of Open Access Journals (Sweden)

    Eduardo Moreira Vicente

    2015-01-01

    Full Text Available This paper presents a new maximum power point tracking (MPPT method based on the measurement of temperature and short-circuit current, in a simple and efficient approach. These measurements, which can precisely define the maximum power point (MPP, have not been used together in other existing techniques. The temperature is measured with a low cost sensor and the solar irradiance is estimated through the relationship of the measured short-circuit current and its reference. Fast tracking speed and stable steady-state operation are advantages of this technique, which presents higher performance when compared to other well-known techniques.

  18. Continuous temperature measurements on the pouring stand for casting moulds

    Directory of Open Access Journals (Sweden)

    W. Leśniewski

    2008-04-01

    Full Text Available The results of temperature measurements of liquid iron alloys obtained by means of the pyrometer, PDR-1800 series, are presented in the paper. The measurements were performed in conditions determined by the kind of a pouring device. The results obtained for bottom-tap ladles were supplemented by laboratory measurements. These results allow explaining significant differences in the results of temperature measurements performed in pouring ladles by means of the pyrometric method and immersible thermocouple, which - in turn - improves assessment of metal thermal parameters in pouring devices.

  19. Low temperature plasma technology methods and applications

    CERN Document Server

    Chu, Paul K

    2013-01-01

    Written by a team of pioneering scientists from around the world, Low Temperature Plasma Technology: Methods and Applications brings together recent technological advances and research in the rapidly growing field of low temperature plasmas. The book provides a comprehensive overview of related phenomena such as plasma bullets, plasma penetration into biofilms, discharge-mode transition of atmospheric pressure plasmas, and self-organization of microdischarges. It describes relevant technology and diagnostics, including nanosecond pulsed discharge, cavity ringdown spectroscopy, and laser-induce

  20. Plasma temperature measurements in disruption simulated experiment

    Energy Technology Data Exchange (ETDEWEB)

    Arkhipov, N.I. [Troitsk Inst. for Innovation and Fusion Research (Russian Federation); Bakhtin, V.P. [Troitsk Inst. for Innovation and Fusion Research (Russian Federation); Safronov, V.M. [Troitsk Inst. for Innovation and Fusion Research (Russian Federation); Toporkov, D.A. [Troitsk Inst. for Innovation and Fusion Research (Russian Federation); Vasenin, S.G. [Troitsk Inst. for Innovation and Fusion Research (Russian Federation); Wurz, H. [Kernforschungszentrum Karlsruhe, INR (Germany); Zhitlukhin, A.M. [Troitsk Inst. for Innovation and Fusion Research (Russian Federation)

    1995-12-31

    Results are reported of experiments to measure the temporal and spatial distributions of a temperature and radiation of a near surface plasma cloud appearing in the disruption simulated experiments. These measurements are needed to verificate the different numerical models of vapor shielding layer which appears to arise near the divertor plates surface and prevents them from the bulk of the incoming energy. Experiments with graphite and tungsten samples were carried out at the 2MK-200 plasma facility. Long CUSP trap was used as a source of high temperature deuterium plasma with a power density W = 10 MW/cm{sup 2} and time duration t = 20 mcs. Laser scattering, space and time resolved soft x-ray spectroscopy was employed to measure the plasma cloud temperature and radiation. The different behaviour of shielding layer parameters was shown for a graphite and tungsten samples. For a tungsten the sharp boundary existed between the incoming deuterium plasma and the thin layer of ablated material plasma and the strong gradient of electron temperature took place in this zone. For a graphite this boundary was broadened at the distance and the main part of the screening layer consisted of the mixture of the incoming deuterium and ablated carbon plasma. (orig.).

  1. Dielectric properties measurement system at cryogenic temperatures and microwave frequencies

    Energy Technology Data Exchange (ETDEWEB)

    Molla, J.; Ibarra, A.; Margineda, J.; Zamarro, J. M.; Hernandez, A.

    1994-07-01

    A system based on the resonant cavity method has been developed to measure the permittivity and loss tangent at 12-18 GHz over the temperature range 80 K to 300 K. Changes of permittivity as low as 0.01 % in the range 1 to 30, and 3 x 10{sup 6} for loss tangent values below 10{sup 2}, can be obtained without requiring temperature stability. The thermal expansion coefficient and resistivity factor of copper have been measured between 80 K and 300 K. Permittivity of sapphire and loss tangent of alumina of 99.9 % purity in the same temperature range are presented. (Author) 23 refs.

  2. Temperature scaling method for Markov chains.

    Science.gov (United States)

    Crosby, Lonnie D; Windus, Theresa L

    2009-01-22

    The use of ab initio potentials in Monte Carlo simulations aimed at investigating the nucleation kinetics of water clusters is complicated by the computational expense of the potential energy determinations. Furthermore, the common desire to investigate the temperature dependence of kinetic properties leads to an urgent need to reduce the expense of performing simulations at many different temperatures. A method is detailed that allows a Markov chain (obtained via Monte Carlo) at one temperature to be scaled to other temperatures of interest without the need to perform additional large simulations. This Markov chain temperature-scaling (TeS) can be generally applied to simulations geared for numerous applications. This paper shows the quality of results which can be obtained by TeS and the possible quantities which may be extracted from scaled Markov chains. Results are obtained for a 1-D analytical potential for which the exact solutions are known. Also, this method is applied to water clusters consisting of between 2 and 5 monomers, using Dynamical Nucleation Theory to determine the evaporation rate constant for monomer loss. Although ab initio potentials are not utilized in this paper, the benefit of this method is made apparent by using the Dang-Chang polarizable classical potential for water to obtain statistical properties at various temperatures.

  3. Novel silver-tubing method for quantitative introduction of water into high-temperature conversion systems for stable hydrogen and oxygen isotopic measurements.

    Science.gov (United States)

    Qi, Haiping; Gröning, Manfred; Coplen, Tyler B; Buck, Bryan; Mroczkowski, Stanley J; Brand, Willi A; Geilmann, Heike; Gehre, Matthias

    2010-07-15

    A new method to seal water in silver tubes for use in a TC/EA (thermal conversion/elemental analyzer) reduction unit using a semi-automated sealing apparatus can yield reproducibilities (1 standard deviation) of delta(2)H and delta(18)O measurements of 1.0 per thousand and 0.06 per thousand, respectively. These silver tubes containing reference waters may be preferred for the calibration of H- and O-bearing materials analyzed with a TC/EA reduction unit. The new sealing apparatus employs a computer-controlled stepping motor to produce silver tubes identical in length. The reproducibility of the mass of water sealed in tubes (in a range of 200-400 microg) can be as good as 1%. Approximately 99% of the sealed silver tubes are satisfactory (leak free). Although silver tubes sealed with reference waters are robust and can be shaken or heated to 110 degrees C with no loss of integrity, they should not be frozen because the expansion during the phase transition of water to ice will break the cold seals and all the water will be lost. The tubes should be shipped in insulated containers. This new method eliminates air inclusions and isotopic fractionation of water associated with the loading of water into capsules using a syringe. The method is also more than an order of magnitude faster than preparing water samples in ordinary Ag capsules. Nevertheless, some laboratories may prefer loading water into silver capsules because expensive equipment is not needed, but users of this method are cautioned to apply the necessary corrections for evaporation, back exchange with laboratory atmospheric moisture, and blanks. Copyright 2010 John Wiley & Sons, Ltd.

  4. The realization of temperature controller for small resistance measurement system

    Science.gov (United States)

    Sobecki, Jakub; Walendziuk, Wojciech; Idzkowski, Adam

    2017-08-01

    This paper concerns the issues of construction and experimental tests of a temperature stabilization system for small resistance increments measurement circuits. After switching the system on, a PCB board heats up and the long-term temperature drift altered the measurement result. The aim of this work is reducing the time of achieving constant nominal temperature by the measurement system, which would enable decreasing the time of measurements in the steady state. Moreover, the influence of temperatures higher than the nominal on the measurement results and the obtained heating curve were tested. During the working process, the circuit heats up to about 32 °C spontaneously, and it has the time to reach steady state of about 1200 s. Implementing a USART terminal on the PC and an NI USB-6341 data acquisition card makes recording the data (concerning temperature and resistance) in the digital form and its further processing easier. It also enables changing the quantity of the regulator settings. This paper presents sample results of measurements for several temperature values and the characteristics of the temperature and resistance changes in time as well as their comparison with the output values. The object identification is accomplished due to the Ziegler-Nichols method. The algorithm of determining the step characteristics parameters and examples of computations of the regulator settings are included together with example characteristics of the object regulation.

  5. Continuous Emission Spectrum Measurement for Electron Temperature Determination in Low-Temperature Collisional Plasmas

    International Nuclear Information System (INIS)

    Liu Qiuyan; Li Hong; Chen Zhipeng; Xie Jinlin; Liu Wandong

    2011-01-01

    Continuous emission spectrum measurement is applied for the inconvenient diagnostics of low-temperature collisional plasmas. According to the physical mechanism of continuous emission, a simplified model is presented to analyze the spectrum in low temperature plasma. The validity of this model is discussed in a wide range of discharge parameters, including electron temperature and ionization degree. Through the simplified model, the continuous emission spectrum in a collisional argon internal inductively coupled plasma is experimentally measured to determine the electron temperature distribution for different gas pressures and radio-frequency powers. The inverse Abel transform is also applied for a better spatially resoluted results. Meanwhile, the result of the continuous emission spectrum measurement is compared to that of the electrostatic double probes, which indicates the effectiveness of this method. (low temperature plasma)

  6. Thermal conductivity measurements at cryogenic temperatures at LASA

    International Nuclear Information System (INIS)

    Broggi, F.; Pedrini, D.; Rossi, L.

    1995-08-01

    Here the improvement realised to have better control of the reference junction temperature and measurements carried out on Nb 3 Sn cut out from 2 different coils (named LASA3 and LASA5), showing the difference between the longitudinal and the transverse thermal conductivity, is described. Two different methods of data analysis are presented, the DAM (derivative approximated method) and the TCI (thermal conductivity integral. The data analysis for the tungsten and the LASA5 coil has been done according to the two methods showing that the TCI method with polynomial functions is not adequate to describe the thermal conductivity. Only a polynomial fit based on the TCI method but limited at a lower order than the nominal, when the data are well distributed along the range of measurements, can describe reasonably the thermal conductivity dependence with the temperature. Finally the measurements on a rod of BSCCO 2212 high T c superconductor are presented

  7. Global rainbow refractometry for droplet temperature measurement

    Energy Technology Data Exchange (ETDEWEB)

    Pascal Lemaitre; Emmanuel Porcheron; Amandine Nuboer; Philippe Brun; Pierre Cornet; Jeanne Malet; Jacques Vendel; Laurent Bouilloux [Institut de Radioprotection et de Surete Nucleaire DSU/SERAC, BP 68, 91192 Gif-sur-Yvette Cedex (France); Gerard Grehan [UMR 6614 CORIA, Laboratoire d' Electromagnetisme et Systemes Particulaires Site Universitaire du Madrillet, Avenue de l' universite BP 12, 76 801 Saint Etienne du Rouvray Cedex, (France)

    2005-07-01

    Full text of publication follows: In order to establish an accurate database to characterize the heat and mass transfers between a spray and the atmosphere with thermal-hydraulic conditions representative of a hypothetical nuclear reactor accident in the containment enclosure of a pressurized water reactor (PWR), the French Institut de Radioprotection et de Surete Nucleaire (IRSN) has developed the TOSQAN experimental facility. This experiment is highly instrumented with non-intrusive diagnostics allowing to measure droplet size and velocity and gas concentrations [1]. The aim of this work is to present the Global Rainbow Thermometry (GRT), which is an advanced non-intrusive optical diagnostic, developed to measure the mean temperature of a set of falling droplets, in a measurement volume of 1 cm{sup 3}. The final paper will be divided in three parts. In the first one, we will explain the principle of the rainbow formation and how droplet temperature can be deduced from the rainbow analysis [2]. This part will be illustrated with the theoretical background on the rainbow and numerical simulations of the global rainbow. The second part will be devoted to present the global rainbow experimental set-up we have developed on optical table, its experimental qualification and finally its implementation on the TOSQAN facility [3]. Finally, we will present the temperature measurements achieved in TOSQAN for thermal-hydraulic conditions representative of a hypothetical nuclear reactor accident. These measurements are useful to characterize the heat and mass transfers between the spraying droplets and the air-steam mixture composing the atmosphere. This analysis will be exposed in a two companion papers. References: [1] E. Porcheron, P. Brun, P. Cornet, J. Malet, J. Vendel. Optical diagnostics applied for single and multi-phase flow characterization in the TOSQAN facility dedicated for thermal hydraulic containment studies. NURETH-10 Seoul, Korea, October 5-9, 2003. [2] P

  8. Global rainbow refractometry for droplet temperature measurement

    International Nuclear Information System (INIS)

    Pascal Lemaitre; Emmanuel Porcheron; Amandine Nuboer; Philippe Brun; Pierre Cornet; Jeanne Malet; Jacques Vendel; Laurent Bouilloux; Gerard Grehan

    2005-01-01

    Full text of publication follows: In order to establish an accurate database to characterize the heat and mass transfers between a spray and the atmosphere with thermal-hydraulic conditions representative of a hypothetical nuclear reactor accident in the containment enclosure of a pressurized water reactor (PWR), the French Institut de Radioprotection et de Surete Nucleaire (IRSN) has developed the TOSQAN experimental facility. This experiment is highly instrumented with non-intrusive diagnostics allowing to measure droplet size and velocity and gas concentrations [1]. The aim of this work is to present the Global Rainbow Thermometry (GRT), which is an advanced non-intrusive optical diagnostic, developed to measure the mean temperature of a set of falling droplets, in a measurement volume of 1 cm 3 . The final paper will be divided in three parts. In the first one, we will explain the principle of the rainbow formation and how droplet temperature can be deduced from the rainbow analysis [2]. This part will be illustrated with the theoretical background on the rainbow and numerical simulations of the global rainbow. The second part will be devoted to present the global rainbow experimental set-up we have developed on optical table, its experimental qualification and finally its implementation on the TOSQAN facility [3]. Finally, we will present the temperature measurements achieved in TOSQAN for thermal-hydraulic conditions representative of a hypothetical nuclear reactor accident. These measurements are useful to characterize the heat and mass transfers between the spraying droplets and the air-steam mixture composing the atmosphere. This analysis will be exposed in a two companion papers. References: [1] E. Porcheron, P. Brun, P. Cornet, J. Malet, J. Vendel. Optical diagnostics applied for single and multi-phase flow characterization in the TOSQAN facility dedicated for thermal hydraulic containment studies. NURETH-10 Seoul, Korea, October 5-9, 2003. [2] P

  9. Temperature measurement in French atomic piles

    International Nuclear Information System (INIS)

    Weill, J.; Rastoix, G.

    1950-10-01

    In the Chatillon reactor the temperature is measured (1) in the interior of one of the vertical A1 cylinders filled with UO 2 (temperature interval 20 to 70 deg. C), and (2) in the center of the tank containing D 2 O (20 to 50 deg. C). The instruments used are silver-constantan thermocouples; the wires are insulated by SiO 2 sheaths, those immersed in D 2 O being placed within Al cases 10 mm diameter. In the Saclay reactor the temperature is taken (1) in the interior of 4 U rods (20 to 300 deg. C), (2) at 2 points of the D 2 O mass (20 to 60 deg. C), (3) at one point in graphite (20 to 100 deg. C), and (4) at 5 points in the catalytic setup (200 deg. C). Copper-constantan couples are used (Ag-constantan is not suitable above 150 deg. C); the wires are enclosed in a sheath of glass fabric. In both reactors the accuracy of the temperature measurements is 0.5 deg. C. (author)

  10. Measurements of temperature on LHC thermal models

    CERN Document Server

    Darve, C

    2001-01-01

    Full-scale thermal models for the Large Hadron Collider (LHC) accelerator cryogenic system have been studied at CERN and at Fermilab. Thermal measurements based on two different models permitted us to evaluate the performance of the LHC dipole cryostats as well as to validate the LHC Interaction Region (IR) inner triplet cooling scheme. The experimental procedures made use of temperature sensors supplied by industry and assembled on specially designed supports. The described thermal models took the advantage of advances in cryogenic thermometry which will be implemented in the future LHC accelerator to meet the strict requirements of the LHC for precision, accuracy, reliability, and ease-of-use. The sensors used in the temperature measurement of the superfluid (He II) systems are the primary focus of this paper, although some aspects of the LHC control system and signal conditioning are also reviewed. (15 refs).

  11. Automated measurement of cattle surface temperature and its correlation with rectal temperature.

    Directory of Open Access Journals (Sweden)

    HongXiang Kou

    Full Text Available The body temperature of cattle varies regularly with both the reproductive cycle and disease status. Establishing an automatic method for monitoring body temperature may facilitate better management of reproduction and disease control in cattle. Here, we developed an Automatic Measurement System for Cattle's Surface Temperature (AMSCST to measure the temperature of metatarsus by attaching a special shell designed to fit the anatomy of cattle's hind leg. Using AMSCST, the surface temperature (ST on the metatarsus of the hind leg was successively measured during 24 hours a day with an interval of one hour in three tested seasons. Based on ST and rectal temperature (RT detected by AMSCST and mercury thermometer, respectively, a linear mixed model was established, regarding both the time point and seasonal factors as the fixed effects. Unary linear correlation and Bland-Altman analysis results indicated that the temperatures measured by AMSCST were closely correlated to those measured by mercury thermometer (R2 = 0.998, suggesting that the AMSCST is an accurate and reliable way to detect cattle's body temperature. Statistical analysis showed that the differences of STs among the three seasons, or among the different time points were significant (P<0.05, and the differences of RTs among the different time points were similarly significant (P<0.05. The prediction accuracy of the mixed model was verified by 10-fold cross validation. The average difference between measured RT and predicted RT was about 0.10 ± 0.10°C with the association coefficient of 0.644, indicating the feasibility of this model in measuring cattle body temperature. Therefore, an automated technology for accurately measuring cattle body temperature was accomplished by inventing an optimal device and establishing the AMSCST system.

  12. Sea Surface Temperatures (SST): Significance and Measurement

    Science.gov (United States)

    Singer, S. F.

    2006-05-01

    Oceans cover 71 percent of Earth's surface and control the global climate. Quoted global mean temperature values and trends, largely based on land thermometers, differ substantially -" mainly because of uncertainties about SST. The ongoing controversy about the relative importance of natural climate changes and Anthropogenic Global Warming (AGW) revolves mainly around disparities between temperature trends of the atmosphere and surface (in the tropics and SH, i.e. mostly SST). Accurate measurement of SST is difficult. Geographic coverage is poor and there are many different techniques, each with its own problems and uncertainties: Water temperatures from buckets and ship-engine inlets; fixed and floating buoys; air temperatures from shipboard and island stations; and remote sensing from satellites using IR and microwaves. As is evident, each technique refers to a different level below the air-water interface. Drifter buoys (at around 50 cm) measure temperatures in the euphotic layers that are generally warmer than the bulk mixed layer sampled by ships (typically around 10 m). The IR emission arises from a 10-micron-thick skin that interacts dynamically with the underlying "mixed layer." The microwave data depend also on emissivity and therefore on surface roughness and sea state. SST data derived from corals provide some support for instrumental data but are not conclusive. The majority of corals show a warming trend since 1979; others show cooling or are ambiguous. There are different ways of interpreting this result. Physical optics dictates that the downwelling IR radiation from atmospheric greenhouse gases is absorbed in the first instance within the skin. Only direct measurements can establish how much of this energy is shared with the bulk mixed layer (to which the usual SST values refer.). SST controls evaporation and therefore global precipitation. SST influences tropical cyclones and sea-level rise; but there is lively debate on those issues. Changes in

  13. Temperature measuring element in nuclear reactors

    International Nuclear Information System (INIS)

    Wada, Takashi.

    1987-01-01

    Purpose: To easily measure the partial maximum temperature at a portion within the nuclear reactor where the connection with the external portion is difficult. Constitution: Sodium, potassium or the alloy thereof with high heat expansion coefficient is packed into an elastic vessel having elastic walls contained in a capsule. A piercing member formed into an acute triangle is attached to one end in the direction of expansion and contraction of the elastic container. The two sides of the triangle form an acute knife edge. A diaphragm is disposed within a capsule at a position opposed to the sharpened direction of the piercing member. Such a capsule is placed in a predetermined position of the nuclear reactor. The elastic vessel causes thermal expansion displacement depending on the temperature at a certain position, by which the top end of the pierce member penetrates through the diaphragm. A pierced scar of a penetration length depending on the temperature is resulted to the diaphragm. The length of the piercing damage is electroscopically observed and compared with the calibration curve to recognize the maximum temperature in the predetermined portion of the nuclear reactor. (Kamimura, M.)

  14. Temperature measurements during laser skin welding

    Science.gov (United States)

    Fried, Nathaniel M.; Choi, Bernard; Welch, Ashley J.; Walsh, Joseph T., Jr.

    1999-06-01

    A thermal camera was used to measure surface temperatures during laser skin welding to provide feedback for optimization of the laser parameters. Two-cm-long, full- thickness incisions were made in guinea pig skin. India ink was used as an absorber. Continuous-wave, 1.06-μm, Nd:YAG laser radiation was scanned over the incisions, producing a pulse duration of approximately 100 ms. Cooling durations between scans of 1.6, 4.0, and 8.0 s were studied with total operation times of 3, 5, and 10 min, respectively. A laser spot diameter of 5 mm was used with the power constant at 10 W. Thermal images were obtained at 30 frames per second with a thermal camera detecting 3.5 micrometers radiation. Surface temperatures were recorded at 0, 1, and 6 mm from the center line of the incision. Cooling durations between scans of 1.6 s and 4.0 s in vitro resulted in temperatures at the weld site remaining above ~65°C for prolonged periods of time. Cooling durations between scans as long as 8.0 s were sufficient both in vitro and in vivo to prevent a significant rise in baseline temperatures at the weld site over time.

  15. A nonintrusive temperature measuring system for estimating deep body temperature in bed.

    Science.gov (United States)

    Sim, S Y; Lee, W K; Baek, H J; Park, K S

    2012-01-01

    Deep body temperature is an important indicator that reflects human being's overall physiological states. Existing deep body temperature monitoring systems are too invasive to apply to awake patients for a long time. Therefore, we proposed a nonintrusive deep body temperature measuring system. To estimate deep body temperature nonintrusively, a dual-heat-flux probe and double-sensor probes were embedded in a neck pillow. When a patient uses the neck pillow to rest, the deep body temperature can be assessed using one of the thermometer probes embedded in the neck pillow. We could estimate deep body temperature in 3 different sleep positions. Also, to reduce the initial response time of dual-heat-flux thermometer which measures body temperature in supine position, we employed the curve-fitting method to one subject. And thereby, we could obtain the deep body temperature in a minute. This result shows the possibility that the system can be used as practical temperature monitoring system with appropriate curve-fitting model. In the next study, we would try to establish a general fitting model that can be applied to all of the subjects. In addition, we are planning to extract meaningful health information such as sleep structure analysis from deep body temperature data which are acquired from this system.

  16. Measuring gas temperature during spin-exchange optical pumping process

    Science.gov (United States)

    Normand, E.; Jiang, C. Y.; Brown, D. R.; Robertson, L.; Crow, L.; Tong, X.

    2016-04-01

    The gas temperature inside a Spin-Exchange Optical Pumping (SEOP) laser-pumping polarized 3He cell has long been a mystery. Different experimental methods were employed to measure this temperature but all were based on either modelling or indirect measurement. To date there has not been any direct experimental measurement of this quantity. Here we present the first direct measurement using neutron transmission to accurately determine the number density of 3He, the temperature is obtained using the ideal gas law. Our result showed a surprisingly high gas temperature of 380°C, compared to the 245°C of the 3He cell wall temperature and 178°C of the optical pumping oven temperature. This experiment result may be used to further investigate the unsolved puzzle of the "X-factor" in the SEOP process which places an upper bound to the 3He polarization that can be achieved. Additional spin relaxation mechanisms might exist due to the high gas temperature, which could explain the origin of the X-factor.

  17. Is Oral Temperature an Accurate Measurement of Deep Body Temperature? A Systematic Review

    Science.gov (United States)

    Mazerolle, Stephanie M.; Ganio, Matthew S.; Casa, Douglas J.; Vingren, Jakob; Klau, Jennifer

    2011-01-01

    Context: Oral temperature might not be a valid method to assess core body temperature. However, many clinicians, including athletic trainers, use it rather than criterion standard methods, such as rectal thermometry. Objective: To critically evaluate original research addressing the validity of using oral temperature as a measurement of core body temperature during periods of rest and changing core temperature. Data Sources: In July 2010, we searched the electronic databases PubMed, Scopus, Cumulative Index to Nursing and Allied Health Literature (CINAHL), SPORTDiscus, Academic Search Premier, and the Cochrane Library for the following concepts: core body temperature, oral, and thermometers. Controlled vocabulary was used, when available, as well as key words and variations of those key words. The search was limited to articles focusing on temperature readings and studies involving human participants. Data Synthesis: Original research was reviewed using the Physiotherapy Evidence Database (PEDro). Sixteen studies met the inclusion criteria and subsequently were evaluated by 2 independent reviewers. All 16 were included in the review because they met the minimal PEDro score of 4 points (of 10 possible points), with all but 2 scoring 5 points. A critical review of these studies indicated a disparity between oral and criterion standard temperature methods (eg, rectal and esophageal) specifically as the temperature increased. The difference was −0.50°C ± 0.31°C at rest and −0.58°C ± 0.75°C during a nonsteady state. Conclusions: Evidence suggests that, regardless of whether the assessment is recorded at rest or during periods of changing core temperature, oral temperature is an unsuitable diagnostic tool for determining body temperature because many measures demonstrated differences greater than the predetermined validity threshold of 0.27°C (0.5°F). In addition, the differences were greatest at the highest rectal temperatures. Oral temperature cannot

  18. Photoacoustic temperature measurements for monitoring of thermal therapy

    Science.gov (United States)

    Wang, Shiou-Han; Wei, Chen-Wei; Jee, Shiou-Hwa; Li, Pai-Chi

    2009-02-01

    Plasmonic photothermal therapy is a new cancer thermotherapy method based on surface plasmon resonance of nanoparticles. It is important to measure the temperature during thermotherapy for safety and efficacy. In this study, we apply a photoacoustic (PA) method for real-time, non-invasive temperature measurements. In particular, this method can be effectively combined with a photothermal therapy system that we developed in parallel. The method is based on the fact that the PA pressure amplitude is linearly related to temperature. To explore its potential, a home-made, 20 MHz PA transducer was used, in which an optical fiber was inserted in its center for emitting laser pulses while the PA signal was simultaneously detected. Continuous wave (CW) laser was used to heat the subject, including both phantoms and mice. The temperature of the region of interest was also measured by a fine-needle thermal couple. Results show that the temperature was linearly proportional to the PA signal with good correlation with the CW laser irradiation. The in vivo study also demonstrated potential of this technique.

  19. Infrared radiometric technique in temperature measurement

    Science.gov (United States)

    Glazer, S.; Madding, R.

    1988-01-01

    One class of commercially available imaging infrared radiometers using cooled detectors is sensitive to radiation over the 3 to 12 micron wavelength band. Spectral filters can tailor instrument sensitivity to specific regions where the target exhibits optimum radiance. The broadband spectral response coupled with real time two-dimensional imaging and emittance/background temperature corrections make the instruments useful for remote measurement of surface temperatures from -20 C to +1500 C. Commonly used radiometric techniques and assumptions are discussed, and performance specifications for a typical modern commercial instrument are presented. The potential usefulness of an imaging infrared radiometer in space laboratories is highlighted through examples of research, nondestructive evaluation, safety, and routine maintenance applications. Future improvements in instrument design and application of the radiometric technique are discussed.

  20. Containerless high temperature property measurements by atomic fluorescence

    Science.gov (United States)

    Schiffman, R. A.; Walker, C. A.

    1984-01-01

    Laser induced fluorescence (LIF) techniques for containerless study of high temperature processes and material properties was studied. Gas jet and electromagnetic levitation and electromagnetic and laser heating techniques are used with LIF in earth-based containerless high temperature experiments. Included are the development of an apparatus and its use in the studies of (1) chemical reactions on Al2O3, molybdenum, tungsten and LaB6 specimens, (2) methods for noncontact specimen temperature measurement, (3) levitation jet properties and (4) radiative lifetime and collisional energy transfer rates for electronically excited atoms.

  1. Radon measurements with charcoal canisters temperature and humidity considerations

    Directory of Open Access Journals (Sweden)

    Živanović Miloš Z.

    2016-01-01

    Full Text Available Radon testing by using open-faced charcoal canisters is a cheap and fast screening method. Many laboratories perform the sampling and measurements according to the United States Environmental Protection Agency method - EPA 520. According to this method, no corrections for temperature are applied and corrections for humidity are based on canister mass gain. The EPA method is practiced in the Vinča Institute of Nuclear Sciences with recycled canisters. In the course of measurements, it was established that the mass gain of the recycled canisters differs from mass gain measured by Environmental Protection Agency in an active atmosphere. In order to quantify and correct these discrepancies, in the laboratory, canisters were exposed for periods of 3 and 4 days between February 2015 and December 2015. Temperature and humidity were monitored continuously and mass gain measured. No significant correlation between mass gain and temperature was found. Based on Environmental Protection Agency calibration data, functional dependence of mass gain on humidity was determined, yielding Environmental Protection Agency mass gain curves. The results of mass gain measurements of recycled canisters were plotted against these curves and a discrepancy confirmed. After correcting the independent variable in the curve equation and calculating the corrected mass gain for recycled canisters, the agreement between measured mass gain and Environmental Protection Agency mass gain curves was attained. [Projekat Ministarstva nauke Republike Srbije, br. III43009: New Technologies for Monitoring and Protection of Environment from Harmful Chemical Substances and Radiation Impact

  2. Fuel centerline temperature measurement experiment in JMTR, 2

    International Nuclear Information System (INIS)

    Ando, Hiroei; Kawamura, Hiroshi; Sezaki, Katsuji; Komukai, Bunsaku

    1980-11-01

    Fuel centerline temperature measurement experiment which is the most fundamental for the LWR fuel safety study, is planned to conduct in JMTR using OWL-1 loop facility. Irradiation of the first test assembly was completed. In this paper, the comparison between measured fuel centerline temperature data and predicted ones by JAERI's FREG-4 code which is a computer program to calculate fuel temperature distribution is made. Furthermore, the data analysis method such as how to estimate local linear power and inpile behavior of the instrumentations are described. The maximum fuel center temperature was 1250 0 C at steady state, the maximum linear power was 320 W/cm, and the maximum burnup was about 1600 MWD/T. (author)

  3. Prediction and measurement of selected phase transformation temperatures of steels

    Directory of Open Access Journals (Sweden)

    Martiník O.

    2017-01-01

    Full Text Available The study deals with precise determination of phase transformation temperatures of steel. A series of experimental measurements were carried out by Differential Thermal Analysis (DTA and Direct Thermal Analysis (TA to obtain temperatures very close to the equilibrium temperatures. There are presented results from the high temperatures region, above 1000°C, with focus on the solidus temperatures (TS, peritectic transition (TP and liquidus (TL of multicomponent steels. The data obtained were verified by statistical evaluation and compared with computational thermodynamic and empirical calculations. The calculations were performed using 15 empirical equations obtained by literature research (10 for TL and 5 for TS, as well as by software InterDendritic Solidification (IDS and Thermo-Calc (2015b, TCFE8; TC. It was verified that both thermo-analytical methods used are set correctly; the results are reproducible, comparable and close to equilibrium state.

  4. Real-time temperature field measurement based on acoustic tomography

    International Nuclear Information System (INIS)

    Bao, Yong; Jia, Jiabin; Polydorides, Nick

    2017-01-01

    Acoustic tomography can be used to measure the temperature field from the time-of-flight (TOF). In order to capture real-time temperature field changes and accurately yield quantitative temperature images, two improvements to the conventional acoustic tomography system are studied: simultaneous acoustic transmission and TOF collection along multiple ray paths, and an offline iteration reconstruction algorithm. During system operation, all the acoustic transceivers send modulated and filtered wideband Kasami sequences simultaneously to facilitate fast and accurate TOF measurements using cross-correlation detection. For image reconstruction, the iteration process is separated and executed offline beforehand to shorten computation time for online temperature field reconstruction. The feasibility and effectiveness of the developed methods are validated in the simulation study. The simulation results demonstrate that the proposed method can reduce the processing time per frame from 160 ms to 20 ms, while the reconstruction error remains less than 5%. Hence, the proposed method has great potential in the measurement of rapid temperature change with good temporal and spatial resolution. (paper)

  5. Apparatus for accurately measuring high temperatures

    Science.gov (United States)

    Smith, D.D.

    The present invention is a thermometer used for measuring furnace temperatures in the range of about 1800/sup 0/ to 2700/sup 0/C. The thermometer comprises a broadband multicolor thermal radiation sensor positioned to be in optical alignment with the end of a blackbody sight tube extending into the furnace. A valve-shutter arrangement is positioned between the radiation sensor and the sight tube and a chamber for containing a charge of high pressure gas is positioned between the valve-shutter arrangement and the radiation sensor. A momentary opening of the valve shutter arrangement allows a pulse of the high gas to purge the sight tube of air-borne thermal radiation contaminants which permits the radiation sensor to accurately measure the thermal radiation emanating from the end of the sight tube.

  6. Reconstructing bottom water temperatures from measurements of temperature and thermal diffusivity in marine sediments

    Science.gov (United States)

    Miesner, F.; Lechleiter, A.; Müller, C.

    2015-07-01

    Continuous monitoring of oceanic bottom water temperatures is a complicated task, even in relatively easy-to-access basins like the North or Baltic seas. Here, a method to determine annual bottom water temperature variations from inverse modeling of instantaneous measurements of temperatures and sediment thermal properties is presented. This concept is similar to climate reconstructions over several thousand years from deep borehole data. However, in contrast, the presented method aims at reconstructing the recent temperature history of the last year from sediment thermal properties and temperatures from only a few meters depth. For solving the heat equation, a commonly used forward model is introduced and analyzed: knowing the bottom water temperature variations for the preceding years and the thermal properties of the sediments, the forward model determines the sediment temperature field. The bottom water temperature variation is modeled as an annual cosine defined by the mean temperature, the amplitude and a phase shift. As the forward model operator is non-linear but low-dimensional, common inversion schemes such as the Newton algorithm can be utilized. The algorithms are tested for artificial data with different noise levels and for two measured data sets: from the North Sea and from the Davis Strait. Both algorithms used show stable and satisfying results with reconstruction errors in the same magnitude as the initial data error. In particular, the artificial data sets are reproduced with accuracy within the bounds of the artificial noise level. Furthermore, the results for the measured North Sea data show small variances and resemble the bottom water temperature variations recorded from a nearby monitoring site with relative errors smaller than 1 % in all parameters.

  7. Ultra sound absorption measurements in rock samples at low temperatures

    Science.gov (United States)

    Herminghaus, C.; Berckhemer, H.

    1974-01-01

    A new technique, comparable with the reverberation method in room acoustics, is described. It allows Q-measurements at rock samples of arbitrary shape in the frequency range of 50 to 600 kHz in vacuum (.1 mtorr) and at low temperatures (+20 to -180 C). The method was developed in particular to investigate rock samples under lunar conditions. Ultrasound absorption has been measured at volcanics, breccia, gabbros, feldspar and quartz of different grain size and texture yielding the following results: evacuation raises Q mainly through lowering the humidity in the rock. In a dry compact rock, the effect of evacuation is small. With decreasing temperature, Q generally increases. Between +20 and -30 C, Q does not change much. With further decrease of temperature in many cases distinct anomalies appear, where Q becomes frequency dependent.

  8. Innovative Instrumentation and Analysis of the Temperature Measurement for High Temperature Gasification

    Energy Technology Data Exchange (ETDEWEB)

    Seong W. Lee

    2006-09-30

    The project entitled, ''Innovative Instrumentation and Analysis of the Temperature Measurement for High Temperature Gasification'', was successfully completed by the Principal Investigator, Dr. S. Lee and his research team in the Center for Advanced Energy Systems and Environmental Control Technologies at Morgan State University. The major results and outcomes were presented in semi-annual progress reports and annual project review meetings/presentations. Specifically, the literature survey including the gasifier temperature measurement, the ultrasonic application in cleaning application, and spray coating process and the gasifier simulator (cold model) testing has been successfully conducted during the first year. The results show that four factors (blower voltage, ultrasonic application, injection time intervals, particle weight) were considered as significant factors that affect the temperature measurement. Then the gasifier simulator (hot model) design and the fabrication as well as the systematic tests on hot model were completed to test the significant factors on temperature measurement in the second year. The advanced Industrial analytic methods such as statistics-based experimental design, analysis of variance (ANOVA) and regression methods were applied in the hot model tests. The results show that operational parameters (i.e. air flow rate, water flow rate, fine dust particle amount, ammonia addition) presented significant impact on the temperature measurement inside the gasifier simulator. The experimental design and ANOVA are very efficient way to design and analyze the experiments. The results show that the air flow rate and fine dust particle amount are statistically significant to the temperature measurement. The regression model provided the functional relation between the temperature and these factors with substantial accuracy. In the last year of the project period, the ultrasonic and subsonic cleaning methods and coating

  9. Improved PID method of temperature control for adiabatic demagnetization refrigerators

    International Nuclear Information System (INIS)

    Hoshino, A.; Shinozaki, K.; Ishisaki, Y.; Mihara, T.

    2006-01-01

    We report a new method of precise temperature control for an adiabatic demagnetization refrigerator (ADR). Temperature of the experimental stage of ADRs is usually controlled with the standard PID (Proportional, Integral, and Derivative control) method by decreasing the magnet current of the superconducting solenoid surrounding the paramagnetic salt inside the ADR. In controlling the temperature of our portable ADR system, we found a small residual between the aimed and measured temperatures, which gradually increased in time as the magnet current decreases. This phenomenon is explained by the magnetic cooling theory, and we have introduced a new functional parameter to improve the standard PID method. Applying this improvement to our system, highly stabilized temperature of 10μK rms at 100mK up to the period of ∼15ks is presented. It is demonstrated that the temperature controlled time was increased by ∼30% in our experiment. Our improved PID method is useful to maintain the long-term temperature stability down to almost zero magnet current with a relatively small ADR

  10. SAR measurement in MRI: an improved method

    Science.gov (United States)

    Romano, Rocco; Acernese, Fausto; Indovina, Pietro Luigi; Barone, Fabrizio

    2009-03-01

    During an MR procedure, the patient absorbs a portion of the transmitted RF energy, which may result in tissue heating and other adverse effects, such as alterations in visual, auditory and neural functions. The Specific Absorption Rate (SAR), in W/kg, is the RF power absorbed per unit mass of tissue and is one of the most important parameters related with thermal effects and acts as a guideline for MRI safety. Strict limits to the SAR levels are imposed by patient safety international regulations (CEI - EN 60601 - 2 - 33) and SAR measurements are required in order to verify its respect. The recommended methods for mean SAR measurement are quite problematic and often require a maintenance man intervention and long stop machine. For example, in the CEI recommended pulse energy method, the presence of a maintenance man is required in order to correctly connect the required instrumentation; furthermore, the procedure is complex and requires remarkable processing and calculus. Simpler are the calorimetric methods, also if in this case long acquisition times are required in order to have significant temperature variations and accurate heat capacity knowledge (CEI - EN 60601 - 2- 33). The phase transition method is a new method to measure SAR in MRI which has the advantages to be very simple and to overcome all the typical calorimetric method problems. It does not require in gantry temperature measurements, any specific heat or heat capacity knowledge, but only mass and time measurement. Furthermore, in this method, it is possible to show that all deposited SAR power can be considered acquired and measured.

  11. Refractive index measurement based on confocal method

    Science.gov (United States)

    An, Zhe; Xu, XiPing; Yang, JinHua; Qiao, Yang; Liu, Yang

    2017-10-01

    The development of transparent materials is closed to optoelectronic technology. It plays an increasingly important role in various fields. It is not only widely used in optical lens, optical element, optical fiber grating, optoelectronics, but also widely used in the building material, pharmaceutical industry with vessel, aircraft windshield and daily wear glasses.Regard of solving the problem of refractive index measurement in optical transparent materials. We proposed that using the polychromatic confocal method to measuring the refractive index of transparent materials. In this article, we describes the principle of polychromatic confocal method for measuring the refractive index of glass,and sketched the optical system and its optimization. Then we establish the measurement model of the refractive index, and set up the experimental system. In this way, the refractive index of the glass has been calibrated for refractive index experiment. Due to the error in the experimental process, we manipulated the experiment data to compensate the refractive index measurement formula. The experiment taking the quartz glass for instance. The measurement accuracy of the refractive index of the glass is +/-1.8×10-5. This method is more practical and accurate, especially suitable for non-contact measurement occasions, which environmental requirements is not high. Environmental requirements are not high, the ordinary glass production line up to the ambient temperature can be fully adapted. There is no need for the color of the measured object that you can measure the white and a variety of colored glass.

  12. Determining noncondensible gas fractions at elevated temperatures and pressures using wet and dry bulb temperature measurements

    International Nuclear Information System (INIS)

    Griffith, P.; Bowman, J.

    1987-01-01

    The work reported in this note was undertaken to provide a method of determining the noncondensible gas fractions in a steam-gas mixture such as might be found in large reactor safety experiment like LOFT. In essence, the method used involves measuring the wet and dry bulb temperatures and using an algorithm, in place of the psychometric chart, to determine the partial pressure of the noncondensible gas in the mixture. In accomplishing this, the authors did the following: (1) extended the use of wet and dry-bulb temperature readings to determine mixture composition up to a temperature of 589 K and a pressure of 4.13 x 10 6 Pa. (2) developed an algorithm to reduce the data (3) found which materials would survive those temperatures

  13. Temperature Measurements of Dense Plasmas by Detailed Balance

    International Nuclear Information System (INIS)

    Holl, A; Redmer, R; Ropke, G; Reinholz, H; Thiele, R; Fortmann, C; Forster, E; Cao, L; Tschentscher, T; Toleikis, S; Glenzer, S H

    2006-01-01

    Plasmas at high electron densities of n e = 10 20 - 10 26 cm -3 and moderate temperatures T e = 1 - 20 eV are important for laboratory astrophysics, high energy density science and inertial confinement fusion. These plasmas are usually referred to as Warm Dense Matter (WDM) and are characterized by a coupling parameter of Λ ∼> 1 where correlations become important. The characterization of such plasmas is still a challenging task due to the lack of direct measurement techniques for temperatures and densities. They propose to measure the Thomson scattering spectrum of vacuum-UV radiation off density fluctuations in the plasma. Collective Thomson scattering provides accurate data for the electron temperature applying first principles. Further, this method takes advantage of the spectral asymmetry resulting from detailed balance and is independent of collisional effects in these dense systems

  14. Actual evaporation estimation from infrared measurement of soil surface temperature

    Directory of Open Access Journals (Sweden)

    Davide Pognant

    2013-09-01

    Full Text Available Within the hydrological cycle, actual evaporation represents the second most important process in terms of volumes of water transported, second only to the precipitation phenomena. Several methods for the estimation of the Ea were proposed by researchers in scientific literature, but the estimation of the Ea from potential evapotranspiration often requires the knowledge of hard-to-find parameters (e.g.: vegetation morphology, vegetation cover, interception of rainfall by the canopy, evaporation from the canopy surface and uptake of water by plant roots and many existing database are characterized by missing or incomplete information that leads to a rough estimation of the actual evaporation amount. Starting from the above considerations, the aim of this study is to develop and validate a method for the estimation of the Ea based on two steps: i the potential evaporation estimation by using the meteorological data (i.e. Penman-Monteith; ii application of a correction factor based on the infrared soil surface temperature measurements. The dataset used in this study were collected during two measurement campaigns conducted both in a plain testing site (Grugliasco, Italy, and in a mountain South-East facing slope (Cogne, Italy. During those periods, hourly measurement of air temperature, wind speed, infrared surface temperature, soil heat flux, and soil water content were collected. Results from the dataset collected in the two testing sites show a good agreement between the proposed method and reference methods used for the Ea estimation.

  15. Crowdsourcing urban air temperature measurements using smartphones

    Science.gov (United States)

    Balcerak, Ernie

    2013-10-01

    Crowdsourced data from cell phone battery temperature sensors could be used to contribute to improved real-time, high-resolution air temperature estimates in urban areas, a new study shows. Temperature observations in cities are in some cases currently limited to a few weather stations, but there are millions of smartphone users in many cities. The batteries in cell phones have temperature sensors to avoid damage to the phone.

  16. Electron temperature measurement of tungsten inert gas arcs

    International Nuclear Information System (INIS)

    Tanaka, Manabu; Tashiro, Shinichi

    2008-01-01

    In order to make clear the physical grounds of deviations from LTE (Local Thermodynamic Equilibrium) in the atmospheric helium TIG arcs electron temperature and LTE temperature obtained from electron number density were measured by using of line-profile analysis of the laser scattering method without an assumption of LTE. The experimental results showed that in comparison with the argon TIG arcs, the region where a deviation from LTE occurs tends to expand in higher arc current because the plasma reaches the similar state to LTE within shorter distance from the cathode due to the slower cathode jet velocity

  17. Thin-film resistance temperature detector array for the measurement of temperature distribution inside a phantom

    Science.gov (United States)

    Sim, Jai Kyoung; Hyun, Jaeyub; Doh, Il; Ahn, Bongyoung; Kim, Yong Tae

    2018-02-01

    A thin-film resistance temperature detector (RTD) array is proposed to measure the temperature distribution inside a phantom. HIFU (high-intensity focused ultrasound) is a non-invasive treatment method using focused ultrasound to heat up a localized region, so it is important to measure the temperature distribution without affecting the ultrasonic field and heat conduction. The present 25 µm thick PI (polyimide) film is transparent not only to an ultrasonic field, because its thickness is much smaller than the wavelength of ultrasound, but also to heat conduction, owing to its negligible thermal mass compared to the phantom. A total of 33 RTDs consisting of Pt resistors and interconnection lines were patterned on a PI substrate using MEMS (microelectromechanical systems) technology, and a polymer phantom was fabricated with the film at the center. The expanded uncertainty of the RTDs was 0.8 K. In the experimental study using a 1 MHz HIFU transducer, the maximum temperature inside the phantom was measured as 70.1 °C just after a HIFU excitation of 6.4 W for 180 s. The time responses of the RTDs at different positions also showed the residual heat transfer inside the phantom after HIFU excitation. HIFU results with the phantom showed that a thin-film RTD array can measure the temperature distribution inside a phantom.

  18. Automated Methods of Corrosion Measurements

    DEFF Research Database (Denmark)

    Andersen, Jens Enevold Thaulov

    1997-01-01

    Scanning probe microscopy (SPM) techniques rely on computer recordings of interactions between the tip of a minute probe and the surface of the small specimen as a function of position; the measurements are used to depict an image of the atomic-scale surface topography on the computer screen....... Mechanical control, recording, and data processing must therefore be automated to a high level of precision and reliability. These general techniques and the apparatus involved have been described extensively. The automated methods of such high-resolution microscopy coordinated with computerized...... electrochemical measurements as well as elemental analysis look very promising for elucidating corrosion reaction mechanisms. The study of initial surface reactions at the atomic or submicron level is becoming an important field of research in the understanding of corrosion processes. At present, mainly two...

  19. Elevated-temperature luminescence measurements to improve spatial resolution

    Science.gov (United States)

    Pluska, Mariusz; Czerwinski, Andrzej

    2018-01-01

    Various branches of applied physics use luminescence based methods to investigate light-emitting specimens with high spatial resolution. A key problem is that luminescence signals lack all the advantages of high locality (i.e. of high spatial resolution) when structures with strong built-in electric field are measured. Such fields exist intentionally in most photonic structures, and occur unintentionally in many other materials. In this case, as a result of beam-induced current generation and its outflow, information that indicates irregularities, nonuniformities and inhomogeneities, such as defects, is lost. We show that to avoid nonlocality and enable truly local luminescence measurements, an elevated measurement temperature as high as 350 K (or even higher) is, perhaps surprisingly, advantageous. This is in contrast to a widely used approach, where cryogenic temperatures, or at least room temperature, are recommended. The elevated temperature of a specimen, together with the current outflow being limited by focused ion beam (FIB) milling, is shown to improve the spatial resolution of luminescence measurements greatly. All conclusions drawn using the example of cathodoluminescence are useful for other luminescence techniques.

  20. Two-Color Pyrometer for Process Temperature Measurement During Machining

    OpenAIRE

    Tapetado Moraleda, Alberto; Díaz Álvarez, José; Miguélez Garrido, María Henar; Vázquez García, María Carmen

    2016-01-01

    A fast fiber-optic two-color pyrometer operating on the optical communication bands is designed for temperature measurements in machining processes. Off-the-shelf low-loss fiber-optic demultiplexers and optoelectronics equipment are used in order to obtain a cost-effective sensing solution while reducing both the temperature measurement error and the minimum measurable temperature. The system is capable of measuring highly localized temperatures without using collimation lens. The designed py...

  1. Low temperature fiber optic pyrometer for fast time resolved temperature measurements

    Science.gov (United States)

    Willsch, M.; Bosselmann, T.; Gaenshirt, D.; Kaiser, J.; Villnow, M.; Banda, M.

    2016-05-01

    Low temperature Pyrometry at temperatures beyond 150°C is limited in the measurement speed due to slow pyroelectric detectors. To detect the circumferential temperature distribution of fast rotating machines a novel Fiber Optical Pyrometer Type is presented here.

  2. Air temperature measurements based on the speed of sound to compensate long distance interferometric measurements

    Directory of Open Access Journals (Sweden)

    Astrua Milena

    2014-01-01

    Full Text Available A method to measure the real time temperature distribution along an interferometer path based on the propagation of acoustic waves is presented. It exploits the high sensitivity of the speed of sound in air to the air temperature. In particular, it takes advantage of a special set-up where the generation of the acoustic waves is synchronous with the amplitude modulation of a laser source. A photodetector converts the laser light to an electronic signal considered as reference, while the incoming acoustic waves are focused on a microphone and generate a second signal. In this condition, the phase difference between the two signals substantially depends on the temperature of the air volume interposed between the sources and the receivers. The comparison with the traditional temperature sensors highlighted the limit of the latter in case of fast temperature variations and the advantage of a measurement integrated along the optical path instead of a sampling measurement. The capability of the acoustic method to compensate the interferometric distance measurements due to air temperature variations has been demonstrated for distances up to 27 m.

  3. Comparison of microstickies measurement methods. Part II, Results and discussion

    Science.gov (United States)

    Mahendra R. Doshi; Angeles Blanco; Carlos Negro; Concepcion Monte; Gilles M. Dorris; Carlos C. Castro; Axel Hamann; R. Daniel Haynes; Carl Houtman; Karen Scallon; Hans-Joachim Putz; Hans Johansson; R. A. Venditti; K. Copeland; H.-M. Chang

    2003-01-01

    In part I of the article we discussed sample preparation procedure and described various methods used for the measurement of microstickies. Some of the important features of different methods are highlighted in Table 1. Temperatures used in the measurement methods vary from room temperature in some cases, 45 °C to 65 °C in other cases. Sample size ranges from as low as...

  4. Bubble measuring instrument and method

    Science.gov (United States)

    Kline-Schoder, Robert (Inventor); Magari, Patrick J. (Inventor)

    2003-01-01

    Method and apparatus are provided for a non-invasive bubble measuring instrument operable for detecting, distinguishing, and counting gaseous embolisms such as bubbles over a selectable range of bubble sizes of interest. A selected measurement volume in which bubbles may be detected is insonified by two distinct frequencies from a pump transducer and an image transducer, respectively. The image transducer frequency is much higher than the pump transducer frequency. The relatively low-frequency pump signal is used to excite bubbles to resonate at a frequency related to their diameter. The image transducer is operated in a pulse-echo mode at a controllable repetition rate that transmits bursts of high-frequency ultrasonic signal to the measurement volume in which bubbles may be detected and then receives the echo. From the echo or received signal, a beat signal related to the repetition rate may be extracted and used to indicate the presence or absence of a resonant bubble. In a preferred embodiment, software control maintains the beat signal at a preselected frequency while varying the pump transducer frequency to excite bubbles of different diameters to resonate depending on the range of bubble diameters selected for investigation.

  5. Bulk temperature measurement in thermally striped pipe flows

    International Nuclear Information System (INIS)

    Lemure, N.; Olvera, J.R.; Ruggles, A.E.

    1995-12-01

    The hot leg flows in some Pressurized Water Reactor (PWR) designs have a temperature distribution across the pipe cross-section. This condition is often referred to as a thermally striped flow. Here, the bulk temperature measurement of pipe flows with thermal striping is explored. An experiment is conducted to examine the feasibility of using temperature measurements on the external surface of the pipe to estimate the bulk temperature of the flow. Simple mixing models are used to characterize the development of the temperature profile in the flow. Simple averaging techniques and Backward Propagating Neural Net are used to predict bulk temperature from the external temperature measurements. Accurate bulk temperatures can be predicted. However, some temperature distributions in the flow effectively mask the bulk temperature from the wall and cause significant error in the bulk temperature predicted using this technique

  6. Measured Performance of a Low Temperature Air Source Heat Pump

    Energy Technology Data Exchange (ETDEWEB)

    Johnson, R. K. [Johnson Research LLC, Pueblo West, CO (United States)

    2013-09-01

    A 4-ton Low Temperature Heat Pump (LTHP) manufactured by Hallowell International was installed in a residence near New Haven, Connecticut and monitored over two winters of operation. After attending to some significant service issues, the heat pump operated as designed. This report should be considered a review of the dual compressor 'boosted heat pump' technology. The Low Temperature Heat Pumpsystem operates with four increasing levels of capacity (heat output) as the outdoor temperature drops. The system was shown to select capacity correctly, supplying the appropriate amount of heat to the house across the full range of outdoor temperatures. The system's Coefficient of Performance (Seasonal COP, or SCOP) over two entire winters was calculated, based on measured data, to be 3.29over the first winter and 2.68 over the second winter. A second seasonal efficiency calculation by a different method yielded a SCOP of 2.78 for the first winter and 2.83 for the second winter. This second seasonal efficiency calculation was determined by comparing measured heat pump energy use to the in situ energy use with resistance heat alone. This method is the ratio of the slopes of thedaily energy use load lines.

  7. Hybrid-type temperature sensor for in situ measurement

    Science.gov (United States)

    Iuchi, Tohru; Hiraka, Kensuke

    2006-11-01

    A hybrid-type surface temperature sensor combines the contact and noncontact methods, which allows us to overcome the shortcomings of both methods. The hybrid-type surface thermometer is composed mainly of two components: a metal film sheet that makes contact with an object and a radiometer that is used to detect the radiance of the rear surface of the metal film, which is actually a modified radiation thermometer. Temperature measurement using the hybrid-type thermometer with a several tens micrometer thick Hastelloy sheet, a highly heat and corrosion resistant alloy, is possible with a systematic error of -0.5K and random errors of ±0.5K, in the temperature range from 900to1000K. This thermometer provides a useful means for calibration of in situ temperature measurement in various processes, especially in the silicon semiconductor industry. This article introduces the basic idea of the hybrid-type surface sensor, presents experimental results and discussions, and finally describes some applications.

  8. Experimental evaluation of IGBT junction temperature measurement via peak gate current

    DEFF Research Database (Denmark)

    Baker, Nick; Munk-Nielsen, Stig; Iannuzzo, Francesco

    2015-01-01

    of the internal gate resistor. In this paper, infra-red measurements are used to evaluate the validity of this method. Temperature measurements are made on IGBTs operating under steady state power dissipation. The peak gate current is found to provide a temperature slightly overestimating the temperature...

  9. Temperature Measurement in WTE Boilers Using Suction Pyrometers

    Directory of Open Access Journals (Sweden)

    Fabio Rinaldi

    2013-11-01

    Full Text Available The temperature of the flue-gas in the post combustion zone of a waste to energy (WTE plant has to be maintained within a fairly narrow range of values, the minimum of which is prescribed by the European Waste Directive 2000/76/CE, whereas the maximum value must be such as to ensure the preservation of the materials and the energy efficiency of the plant. A high degree of accuracy in measuring and controlling the aforementioned temperature is therefore required. In almost the totality of WTE plants this measurement process is carried out by using practical industrial thermometers, such as bare thermocouples and infrared radiation (IR pyrometers, even if affected by different physical contributions which can make the gas temperature measurements incorrect. The objective of this paper is to analyze errors and uncertainties that can arise when using a bare thermocouple or an IR pyrometer in a WTE plant and to provide a method for the in situ calibration of these industrial sensors through the use of suction pyrometers. The paper describes principle of operation, design, and uncertainty contributions of suction pyrometers, it also provides the best estimation of the flue-gas temperature in the post combustion zone of a WTE plant and the estimation of its expanded uncertainty.

  10. Temperature measurement in WTE boilers using suction pyrometers.

    Science.gov (United States)

    Rinaldi, Fabio; Najafi, Behzad

    2013-11-15

    The temperature of the flue-gas in the post combustion zone of a waste to energy (WTE) plant has to be maintained within a fairly narrow range of values, the minimum of which is prescribed by the European Waste Directive 2000/76/CE, whereas the maximum value must be such as to ensure the preservation of the materials and the energy efficiency of the plant. A high degree of accuracy in measuring and controlling the aforementioned temperature is therefore required. In almost the totality of WTE plants this measurement process is carried out by using practical industrial thermometers, such as bare thermocouples and infrared radiation (IR) pyrometers, even if affected by different physical contributions which can make the gas temperature measurements incorrect. The objective of this paper is to analyze errors and uncertainties that can arise when using a bare thermocouple or an IR pyrometer in a WTE plant and to provide a method for the in situ calibration of these industrial sensors through the use of suction pyrometers. The paper describes principle of operation, design, and uncertainty contributions of suction pyrometers, it also provides the best estimation of the flue-gas temperature in the post combustion zone of a WTE plant and the estimation of its expanded uncertainty.

  11. Active silicon x-ray for measuring electron temperature

    International Nuclear Information System (INIS)

    Snider, R.T.

    1994-07-01

    Silicon diodes are commonly used for x-ray measurements in the soft x-ray region between a few hundred ev and 20 keV. Recent work by Cho has shown that the charge collecting region in an underbiased silicon detector is the depletion depth plus some contribution from a region near the depleted region due to charge-diffusion. The depletion depth can be fully characterized as a function of the applied bias voltage and is roughly proportional to the squart root of the bias voltage. We propose a technique to exploit this effect to use the silicon within the detector as an actively controlled x-ray filter. With reasonable silicon manufacturing methods, a silicon diode detector can be constructed in which the sensitivity of the collected charge to the impinging photon energy spectrum can be changed dynamically in the visible to above the 20 keV range. This type of detector could be used to measure the electron temperature in, for example, a tokamak plasma by sweeping the applied bias voltage during a plasma discharge. The detector samples different parts of the energy spectrum during the bias sweep, and the data collected contains enough information to determine the electron temperature. Benefits and limitations of this technique will be discussed along with comparisons to similar methods for measuring electron temperature and other applications of an active silicon x-ray filter

  12. INNOVATIVE INSTRUMENTATION AND ANALYSIS OF THE TEMPERATURE MEASUREMENT FOR HIGH TEMPERATURE GASIFICATION

    Energy Technology Data Exchange (ETDEWEB)

    Seong W. Lee

    2003-09-01

    During this reporting period, the literature survey including the gasifier temperature measurement literature, the ultrasonic application and its background study in cleaning application, and spray coating process are completed. The gasifier simulator (cold model) testing has been successfully conducted. Four factors (blower voltage, ultrasonic application, injection time intervals, particle weight) were considered as significant factors that affect the temperature measurement. The Analysis of Variance (ANOVA) was applied to analyze the test data. The analysis shows that all four factors are significant to the temperature measurements in the gasifier simulator (cold model). The regression analysis for the case with the normalized room temperature shows that linear model fits the temperature data with 82% accuracy (18% error). The regression analysis for the case without the normalized room temperature shows 72.5% accuracy (27.5% error). The nonlinear regression analysis indicates a better fit than that of the linear regression. The nonlinear regression model's accuracy is 88.7% (11.3% error) for normalized room temperature case, which is better than the linear regression analysis. The hot model thermocouple sleeve design and fabrication are completed. The gasifier simulator (hot model) design and the fabrication are completed. The system tests of the gasifier simulator (hot model) have been conducted and some modifications have been made. Based on the system tests and results analysis, the gasifier simulator (hot model) has met the proposed design requirement and the ready for system test. The ultrasonic cleaning method is under evaluation and will be further studied for the gasifier simulator (hot model) application. The progress of this project has been on schedule.

  13. Online junction temperature measurement via internal gate resistance during turn-on

    DEFF Research Database (Denmark)

    Baker, Nick; Munk-Nielsen, Stig; Liserre, Marco

    2014-01-01

    A new method for junction temperature measurement of power semiconductor switches is presented. The measurement exploits the temperature dependent resistance of the temperature sensitive electrical parameter (TSEP): the internal gate resistance. This dependence can be observed during the normal s...... TSEP based measurement methods, primarily being: an absence of any dependence on operating conditions such as load current, and the potential to achieve higher sensitivity (20mV/C or more) than alternative TSEPs....

  14. Design of a temperature measurement and feedback control system based on an improved magnetic nanoparticle thermometer

    Science.gov (United States)

    Du, Zhongzhou; Sun, Yi; Liu, Jie; Su, Rijian; Yang, Ming; Li, Nana; Gan, Yong; Ye, Na

    2018-04-01

    Magnetic fluid hyperthermia, as a novel cancer treatment, requires precise temperature control at 315 K-319 K (42 °C-46 °C). However, the traditional temperature measurement method cannot obtain the real-time temperature in vivo, resulting in a lack of temperature feedback during the heating process. In this study, the feasibility of temperature measurement and feedback control using magnetic nanoparticles is proposed and demonstrated. This technique could be applied in hyperthermia. Specifically, the triangular-wave temperature measurement method is improved by reconstructing the original magnetization response of magnetic nanoparticles based on a digital phase-sensitive detection algorithm. The standard deviation of the temperature in the magnetic nanoparticle thermometer is about 0.1256 K. In experiments, the temperature fluctuation of the temperature measurement and feedback control system using magnetic nanoparticles is less than 0.5 K at the expected temperature of 315 K. This shows the feasibility of the temperature measurement method for temperature control. The method provides a new solution for temperature measurement and feedback control in hyperthermia.

  15. IR Camera Validation of IGBT Junction Temperature Measurement via Peak Gate Current

    DEFF Research Database (Denmark)

    Baker, Nick; Dupont, Laurent; Munk-Nielsen, Stig

    2017-01-01

    Infra-red measurements are used to assess the measurement accuracy of the Peak Gate Current (IGPeak) method for IGBT junction temperature measurement. Single IGBT chips with the gate pad in both the centre and the edge are investigated, along with paralleled chips, as well as chips suffering...... partial bond-wire lift-off. Results are also compared with a traditional electrical temperature measurement method: the voltage drop under low current (VCE(low)). In all cases, the IGPeak method is found to provide a temperature slightly overestimating the temperature of the gate pad. Consequently, both...... the gate pad position and chip temperature distribution influence whether the measurement is representative of the mean junction temperature. These results remain consistent after chips are degraded through bondwire lift-off. In a paralleled IGBT configuration with non-negligible temperature disequilibrium...

  16. Efficiency measurement of hydro machine by thermodynamic method

    Energy Technology Data Exchange (ETDEWEB)

    Patil, Shantaram S.; Kumar, Arun [Indian Institute of Technology Roorkee, Uttarakhand (India). Alternate Hydro Energy Centre], E-mails: patiliitr@gmail.com, aheciitr@gmail.com; Verma, H.K. [Indian Institute of Technology Roorkee, Uttarakhand (India). Dept. of Electrical Engineering], E-mail: hkvfee@gmail.com

    2011-10-15

    The thermodynamic method is an absolute method of measuring hydraulic efficiency of hydraulic machines based on the principle of conservation of energy, i.e. first law of thermodynamics. The efficiency is calculated directly from the specific hydraulic energy and specific mechanical energy equations. The temperature difference between the inlet and outlet of the machine is very small, typically of the order of a few milli-Kelvins. Hence it is a very difficult and critical task to measure the temperature rise accurately. High-resolution high-accuracy data acquisition system along with precision type temperature sensors are essential for temperature measurements. Efficiency measurement using thermodynamic method has been carried out by the authors in laboratory, both on a turbine and a pump, using stable temperature sensors, electronic pressure transmitters and high-precision high-resolution data acquisition system. Thermo-wells are formed at inlet and outlet of the machine under test for simultaneous temperature measurement at both the points. High-precision RTDs of Pt-100 type (class 1/10 DIN, 0.03 deg C accuracy) are used for the measurement of temperatures. The results of efficiency measurement of turbine and pump in laboratory using thermodynamic method have been found to be very satisfactory. The same instrumentation with suitable adaptation can be used for the efficiency measurement on site. (author)

  17. 7 CFR 28.301 - Measurement: humidity; temperature.

    Science.gov (United States)

    2010-01-01

    ... 7 Agriculture 2 2010-01-01 2010-01-01 false Measurement: humidity; temperature. 28.301 Section 28... for Length of Staple § 28.301 Measurement: humidity; temperature. The length of staple of any cotton... its fibers under a relative humidity of the atmosphere of 65 percent and a temperature of 70° F. ...

  18. Temperature measurement on neurological pulse generators during MR scans

    Directory of Open Access Journals (Sweden)

    Alesch François

    2002-09-01

    Full Text Available Abstract According to manufacturers of both magnetic resonance imaging (MRI machines, and implantable neurological pulse generators (IPGs, MRI is contraindicated for patients with IPGs. A major argument for this restriction is the risk to induce heat in the leads due to the electromagnetic field, which could be dangerous for the surrounding brain parenchyma. The temperature change on the surface of the case of an ITREL-III (Medtronic Inc., Minneapolis, MN and the lead tip during MRI was determined. An anatomical realistic and a cubic phantom, filled with phantom material mimicking human tissue, and a typical lead configuration were used to imitate a patient who carries an IPG for deep brain stimulation. The measurements were performed in a 1.5 T and a 3.0 T MRI. 2.1°C temperature increases at the lead tip uncovered the lead tip as the most critical part concerning heating problems in IPGs. Temperature increases in other locations were low compared to the one at the lead tip. The measured temperature increase of 2.1°C can not be considered as harmful to the patient. Comparison with the results of other studies revealed the avoidance of loops as a practical method to reduce heating during MRI procedures.

  19. Measuring the temperature of hot nuclear fragments

    International Nuclear Information System (INIS)

    Wuenschel, S.; Bonasera, A.; May, L.W.; Souliotis, G.A.; Tripathi, R.; Galanopoulos, S.; Kohley, Z.; Hagel, K.; Shetty, D.V.; Huseman, K.; Soisson, S.N.; Stein, B.C.; Yennello, S.J.

    2010-01-01

    A new thermometer based on fragment momentum fluctuations is presented. This thermometer exhibited residual contamination from the collective motion of the fragments along the beam axis. For this reason, the transverse direction has been explored. Additionally, a mass dependence was observed for this thermometer. This mass dependence may be the result of the Fermi momentum of nucleons or the different properties of the fragments (binding energy, spin, etc.) which might be more sensitive to different densities and temperatures of the exploding fragments. We expect some of these aspects to be smaller for protons (and/or neutrons); consequently, the proton transverse momentum fluctuations were used to investigate the temperature dependence of the source.

  20. Temperature measurements on a HSLA-100 steel confinement vessel

    Energy Technology Data Exchange (ETDEWEB)

    Lohsen, R.A.

    1998-05-07

    Temperature measurements have been made on HSLA-100 steel confinement vessel number 6-2-3-1. These measurements are intended to give a view of the vessel temperature response under conditions similar to operational conditions, starting from worst case. The vessel`s temperature must be above the minimum operating temperature when used to contain an explosive event to ensure that the vessel material has the desired crack arrest properties. Several series of temperature measurements have been conducted over 24 and 48 hour periods during February 1998. These tests were intended to demonstrate that after running the heaters in the environmental shelter for some time, (1) the vessel warms up to temperatures well above the minimum operating temperature, (2) that through-thickness temperature gradients are negligible, and (3) that the temperature differences from one part of the vessel to another are small.

  1. Low Temperature Hall Measurements of Neutron Irradiated Silicon Carbide

    National Research Council Canada - National Science Library

    Bonavita, Ange1o

    2004-01-01

    .... No features suggesting annealing were found below a temperature of 340K. Temperature dependant Hall effect measurements were taken over a range of 100K to 340K recording resistivity, carrier densities, and mobility...

  2. Isothermal temperature reactivity coefficient measurement in TRIGA reactor

    International Nuclear Information System (INIS)

    Zagar, T.; Ravnik, M.; Trkov, A.

    2002-01-01

    Direct measurement of an isothermal temperature reactivity coefficient at room temperatures in TRIGA Mark II research reactor at Jozef Stefan Institute in Ljubljana is presented. Temperature reactivity coefficient was measured in the temperature range between 15 o C and 25 o C. All reactivity measurements were performed at almost zero reactor power to reduce or completely eliminate nuclear heating. Slow and steady temperature decrease was controlled using the reactor tank cooling system. In this way the temperatures of fuel, of moderator and of coolant were kept in equilibrium throughout the measurements. It was found out that TRIGA reactor core loaded with standard fuel elements with stainless steel cladding has small positive isothermal temperature reactivity coefficient in this temperature range.(author)

  3. Investigation of temperature correction for tire/pavement noise measurements

    Science.gov (United States)

    2010-11-01

    The Volpe Center Acoustics Facility, in support of the Federal Highway Administration, : investigated the influence of temperature on tire/pavement noise in order to provide guidance on correcting for temperature variations in measured sound levels. ...

  4. Experimental relations between airborne and ground measured wheat canopy temperatures

    Science.gov (United States)

    Millard, J. P.; Reginato, R. J.; Idso, S. B.; Jackson, R. D.; Goettelman, R. C.; Leroy, M. J.

    1980-01-01

    Experiments using ground-based measurements of canopy temperatures have shown that plant temperatures are good indicators of plant water stress, and thus are useful for assessing water requirements and predicting yields. An intensive 23-day airborne- and ground-measurement program was conducted in Phoenix, Arizona in 1977 to compare airborne-acquired wheat canopy temperatures with simultaneous ground measurements. For canopies that covered at least 85 percent of the soil surface, airborne measurements differed from ground measurements of plant temperature by less than 2 C. Regardless of the amount of plant cover, the airborne measurements were virtually identical to ground-nadir measurements, and thus represent a combination of plant temperature and solid background temperature.

  5. Electron Density and Temperature Measurements, and Abundance ...

    Indian Academy of Sciences (India)

    tribpo

    tics—emission lines. Dwivedi, Curdt & Wilhelm (1997, 1999a) carried out an observing sequence based on a theoretical study by Dwivedi & Mohan (1995), with intercombination/forbidden. Ne VI and Mg VI lines, which are formed at essentially the same temperature. (4 × 105 K), according to Arnaud & Rothenflug (1985).

  6. Measuring the temperature history of isochorically heated warm dense metals

    Science.gov (United States)

    McGuffey, Chris; Kim, J.; Park, J.; Moody, J.; Emig, J.; Heeter, B.; Dozieres, M.; Beg, Fn; McLean, Hs

    2017-10-01

    A pump-probe platform has been designed for soft X-ray absorption spectroscopy near edge structure measurements in isochorically heated Al or Cu samples with temperature of 10s to 100s of eV. The method is compatible with dual picosecond-class laser systems and may be used to measure the temperature of the sample heated directly by the pump laser or by a laser-driven proton beam Knowledge of the temperature history of warm dense samples will aid equation of state measurements. First, various low- to mid-Z targets were evaluated for their suitability as continuum X-ray backlighters over the range 200-1800 eV using a 10 J picosecond-class laser with relativistic peak intensity Alloys were found to be more suitable than single-element backlighters. Second, the heated sample package was designed with consideration of target thickness and tamp layers using atomic physics codes. The results of the first demonstration attempts will be presented. This work was supported by the U.S. DOE under Contract No. DE-SC0014600.

  7. The measurement for level of marine high-temperature and high-pressure vessels

    International Nuclear Information System (INIS)

    Lin Jie.

    1986-01-01

    The various error factors in measurement for level of marine high-temperature and high-pressure vessels are anslysed. The measuring method of error self compensation and its simplification for land use are shown

  8. Temperature Measurement and Numerical Prediction in Machining Inconel 718.

    Science.gov (United States)

    Díaz-Álvarez, José; Tapetado, Alberto; Vázquez, Carmen; Miguélez, Henar

    2017-06-30

    Thermal issues are critical when machining Ni-based superalloy components designed for high temperature applications. The low thermal conductivity and extreme strain hardening of this family of materials results in elevated temperatures around the cutting area. This elevated temperature could lead to machining-induced damage such as phase changes and residual stresses, resulting in reduced service life of the component. Measurement of temperature during machining is crucial in order to control the cutting process, avoiding workpiece damage. On the other hand, the development of predictive tools based on numerical models helps in the definition of machining processes and the obtainment of difficult to measure parameters such as the penetration of the heated layer. However, the validation of numerical models strongly depends on the accurate measurement of physical parameters such as temperature, ensuring the calibration of the model. This paper focuses on the measurement and prediction of temperature during the machining of Ni-based superalloys. The temperature sensor was based on a fiber-optic two-color pyrometer developed for localized temperature measurements in turning of Inconel 718. The sensor is capable of measuring temperature in the range of 250 to 1200 °C. Temperature evolution is recorded in a lathe at different feed rates and cutting speeds. Measurements were used to calibrate a simplified numerical model for prediction of temperature fields during turning.

  9. Temperature Measurement and Numerical Prediction in Machining Inconel 718

    Science.gov (United States)

    Tapetado, Alberto; Vázquez, Carmen; Miguélez, Henar

    2017-01-01

    Thermal issues are critical when machining Ni-based superalloy components designed for high temperature applications. The low thermal conductivity and extreme strain hardening of this family of materials results in elevated temperatures around the cutting area. This elevated temperature could lead to machining-induced damage such as phase changes and residual stresses, resulting in reduced service life of the component. Measurement of temperature during machining is crucial in order to control the cutting process, avoiding workpiece damage. On the other hand, the development of predictive tools based on numerical models helps in the definition of machining processes and the obtainment of difficult to measure parameters such as the penetration of the heated layer. However, the validation of numerical models strongly depends on the accurate measurement of physical parameters such as temperature, ensuring the calibration of the model. This paper focuses on the measurement and prediction of temperature during the machining of Ni-based superalloys. The temperature sensor was based on a fiber-optic two-color pyrometer developed for localized temperature measurements in turning of Inconel 718. The sensor is capable of measuring temperature in the range of 250 to 1200 °C. Temperature evolution is recorded in a lathe at different feed rates and cutting speeds. Measurements were used to calibrate a simplified numerical model for prediction of temperature fields during turning. PMID:28665312

  10. Cryogenic temperature measurement for large applications

    CERN Document Server

    Ylöstalo, J; Kyynäräinen, J; Niinikoski, T O; Voutilainen, R

    1996-01-01

    We have developed a resistance thermometry system for the acquisition, control and monitoring of temperature in large-scale cryogenic applications. The resistance of the sensor is converted to a voltage using a self-balancing AC bridge circuit featuring square-wave excitation currents down to 1 nA. The system is easily scalable and includes intelligent features to treat special situations such as magnet quenches differently from normal operation.

  11. Design and Implementation of High Precision Temperature Measurement Unit

    Science.gov (United States)

    Zeng, Xianzhen; Yu, Weiyu; Zhang, Zhijian; Liu, Hancheng

    2018-03-01

    Large-scale neutrino detector requires calibration of photomultiplier tubes (PMT) and electronic system in the detector, performed by plotting the calibration source with a group of designated coordinates in the acrylic sphere. Where the calibration source positioning is based on the principle of ultrasonic ranging, the transmission speed of ultrasonic in liquid scintillator of acrylic sphere is related to temperature. This paper presents a temperature measurement unit based on STM32L031 and single-line bus digital temperature sensor TSic506. The measurement data of the temperature measurement unit can help the ultrasonic ranging to be more accurate. The test results show that the temperature measurement error is within ±0.1°C, which satisfies the requirement of calibration source positioning. Take energy-saving measures, with 3.7V/50mAH lithium battery-powered, the temperature measurement unit can work continuously more than 24 hours.

  12. Method for estimating boiling temperatures of crude oils

    International Nuclear Information System (INIS)

    Jones, R.K.

    1996-01-01

    Evaporation is often the dominant mechanism for mass loss during the first few days following an oil spill. The initial boiling point of the oil and the rate at which the boiling point changes as the oil evaporates are needed to initialize some computer models used in spill response. The lack of available boiling point data often limits the usefulness of these models in actual emergency situations. A new computational method was developed to estimate the temperature at which a crude oil boils as a function of the fraction evaporated using only standard distillation data, which are commonly available. This method employs established thermodynamic rules and approximations, and was designed to be used with automated spill-response models. Comparisons with measurements show a strong correlation between results obtained with this method and measured values

  13. ATTREX-Aircraft_RemoteSensing_Temperature_Measurements

    Data.gov (United States)

    National Aeronautics and Space Administration — This collection consists of the observational data from the Airborne Tropical TRopopause EXperiment (ATTREX) spectral and broad band irradiance measurements covering...

  14. Elevated temperature forming method and preheater apparatus

    Science.gov (United States)

    Krajewski, Paul E; Hammar, Richard Harry; Singh, Jugraj; Cedar, Dennis; Friedman, Peter A; Luo, Yingbing

    2013-06-11

    An elevated temperature forming system in which a sheet metal workpiece is provided in a first stage position of a multi-stage pre-heater, is heated to a first stage temperature lower than a desired pre-heat temperature, is moved to a final stage position where it is heated to a desired final stage temperature, is transferred to a forming press, and is formed by the forming press. The preheater includes upper and lower platens that transfer heat into workpieces disposed between the platens. A shim spaces the upper platen from the lower platen by a distance greater than a thickness of the workpieces to be heated by the platens and less than a distance at which the upper platen would require an undesirably high input of energy to effectively heat the workpiece without being pressed into contact with the workpiece.

  15. Nonequilibrium shock layer temperature profiles from arc jet radiation measurements

    Science.gov (United States)

    Blackwell, Harvel E.; Yuen, Eric; Scott, Carl D.; Arepalli, Sivaram

    1989-01-01

    Shock layer temperature profiles are obtained through analysis of radiation from shock layers produced by a blunt body inserted in arc jet flow. Spectral measurements have been made in a nitrogen flow of 54.4 gm/s at an enthalpy of 8.72 MJ/kg. Vibrational temperatures for N2+ are obtained by matching spectral regions from arc jet spectra with spectra generated using the NEQAIR code. Temperature profiles obtained from the radiation layers show a vibrational temperature higher than the rotational temperature near the front of the shock and both temperatures decrease as the flow approaches the body. The spectral measurements are made and analysis completed for four distances, from the surface of the blunt body. The corresponding shock layer thickness is approximately 3.6 cm. Although the shock layer appears to be in thermal nonequilibrium, the measured rotational temperature approaches the single temperature results of viscous shock layer calculations at this test condition.

  16. The liner brightness temperature measurement by two channel optical pyrometer

    Science.gov (United States)

    Kulish, M. I.; Dudin, S. V.; Ushnurtsev, A. E.; Mintsev, V. B.

    2018-01-01

    Measurability of liner inner surface brightness temperature by two channel optical pyrometer is shown. Liner is compressed by detonation products in large-scale experiment. Absolute radiant intensity values were obtained by measuring optical system channel calibration involving tungsten and xenon radiation sources. Three ways of surface brightness temperature measurement are presented at wavelengths of 620 and 850 nm. Using the developed procedure copper and steel liners behavior (brightness temperature, average speed) under compression by detonation products are evaluated.

  17. Temperature Measurement of a Glass Material Using a Multiwavelength Pyrometer

    Science.gov (United States)

    Ng, Daniel

    1997-01-01

    Temperature measurement of a substance that is transparent using the traditional 1-color, 2-color and other pyrometers has been difficult. The radiation detected by pyrometers do not come from a well defined location in the transparent body. The multiwavelength pyrometer developed at the NASA Lewis Research Center can measure the surface temperature of many materials. We show in this paper that it also measures the surface and a bulk subsurface temperature of transparent materials like glass.

  18. Temperature dependent measurement of internal damping of austenitic stainless steels

    Directory of Open Access Journals (Sweden)

    Oravcová Monika

    2018-01-01

    Full Text Available This article is aimed on the analysis of the internal damping changes of austenitic stainless steels AISI 304, AISI 316L and AISI 316Ti depending from temperature. In experimental measurements only resonance method was used which is based on continuous excitation of oscillations of the specimens and the whole apparatus vibrates at the frequency near to the resonance. Microplastic processes and dissipation of energy within the metals are evaluated and investigated by internal damping measurements. Damping capacity of materials is closely tied to the presence of defects including second phase particles and voids. By measuring the energy dissipation in the material, we can determine the elastic characteristics, Youngs modulus, the level of stress relaxation and many other.

  19. The Measurement and Interpretation of Transformation Temperatures in Nitinol

    Science.gov (United States)

    Duerig, T. W.; Pelton, A. R.; Bhattacharya, K.

    2017-12-01

    A previous paper (Duerig and Bhattacharya in Shap Mem Superelasticity 1:153-161, 2015) introduced several engineering considerations surrounding the R-phase in Nitinol and highlighted a common, if not pervasive, misconception regarding the use of the term Af by the medical device industry. This paper brings additional data to bear on the issue and proposes more accurate terminology. Moreover, a variety of tools are used to establish the forward and reverse stress-temperature phase diagrams for a superelastic wire typical of that used in medical devices. Once established, the two most common methods of measuring transformation temperatures, Differential Scanning Calorimetry and Bend Free Recovery, are tested against the observed behavior. Light is also shed upon the origin of the Clausius-Clapeyron ratio (d σ/d T), the triple point, and why such large variations are reported in superelastic alloys.

  20. The balancing of existing space heating systems using temperature measurements; L'equilibrage des installations de chauffage existantes par mesure des temperatures

    Energy Technology Data Exchange (ETDEWEB)

    Didier, G.

    2004-03-01

    The balancing of space heating systems using the measurements of water temperature or of the ambient temperature is a method used since several decades. However, the results obtained with this method have been very often disappointing. Today, the improvements made in the metrology of surface temperatures allow to reconsider this approach. This technical paper describes: the progresses made in surface temperature measurements, the balancing principle using water temperature measurements, a numerical example and the study of a concrete case (relation between temperature drop and flow rate), and the choice of a balancing method. (J.S.)

  1. Method for determining waveguide temperature for acoustic transceiver used in a gas turbine engine

    Energy Technology Data Exchange (ETDEWEB)

    DeSilva, Upul P.; Claussen, Heiko; Ragunathan, Karthik

    2018-04-17

    A method for determining waveguide temperature for at least one waveguide of a transceiver utilized for generating a temperature map. The transceiver generates an acoustic signal that travels through a measurement space in a hot gas flow path defined by a wall such as in a combustor. The method includes calculating a total time of flight for the acoustic signal and subtracting a waveguide travel time from the total time of flight to obtain a measurement space travel time. A temperature map is calculated based on the measurement space travel time. An estimated wall temperature is obtained from the temperature map. An estimated waveguide temperature is then calculated based on the estimated wall temperature wherein the estimated waveguide temperature is determined without the use of a temperature sensing device.

  2. Solar energy control system. [temperature measurement

    Science.gov (United States)

    Currie, J. R. (Inventor)

    1981-01-01

    A solar energy control system for a hot air type solar energy heating system wherein thermocouples are arranged to sense the temperature of a solar collector, a space to be heated, and a top and bottom of a heat storage unit is disclosed. Pertinent thermocouples are differentially connected together, and these are employed to effect the operation of dampers, a fan, and an auxiliary heat source. In accomplishing this, the differential outputs from the thermocouples are amplified by a single amplifier by multiplexing techniques. Additionally, the amplifier is corrected as to offset by including as one multiplex channel a common reference signal.

  3. Device for the alternative option of temperature measurement

    Science.gov (United States)

    Jargus, Jan; Nedoma, Jan; Fajkus, Marcel; Novak, Martin; Cubik, Jakub; Cvejn, Daniel; Vasinek, Vladimir

    2017-10-01

    Polydimethylsiloxane (PDMS) has good optical properties, and its composition offers the possibility of use in many applications (industry, security device, medicine applications and etc.). We focused on the alternative option of temperature measurement in this article. Our approach is based on measuring changes of chromaticity correlated temperature corresponding to changes in temperature. Described device uses an optical fiber with a defined layer of PDMS and luminophore and we assume that it can find use also in the field of security. The article describes the process of making the prototype of the device and its verification based on laboratory results. The measured temperature depends mainly on the type of optical fiber and the measured temperature range is determined by the thermal resistance of used optical fiber. Using a calibration measurement can determine the value of temperature with an accuracy of +/- 2,5 %.

  4. Temperature Measurement of Ceramic Materials Using a Multiwavelength Pyrometer

    Science.gov (United States)

    Ng, Daniel; Fralick, Gustave

    1999-01-01

    The surface temperatures of several pure ceramic materials (alumina, beryllia, magnesia, yittria and spinel) in the shape of pellets were measured using a multiwavelength pyrometer. In one of the measurements, radiation signal collection is provided simply by an optical fiber. In the other experiments, a 4.75 inch (12 cm) parabolic mirror collects the signal for the spectrometer. Temperature measurement using the traditional one- and two-color pyrometer for these ceramic materials is difficult because of their complex optical properties, such as low emissivity which varies with both temperature and wavelength. In at least one of the materials, yittria, the detected optical emission increased as the temperature was decreased due to such emissivity variation. The reasons for such changes are not known. The multiwavelength pyrometer has demonstrated its ability to measure surface temperatures under such conditions. Platinum electrodes were embedded in the ceramic pellets for resistance measurements as the temperature changed.

  5. LIDAR for atmospheric backscatter and temperature measurements

    Data.gov (United States)

    National Aeronautics and Space Administration — We propose to demonstrate key capabilities of a multifunctional atmospheric lidar. This lidar could be part of a Surface Weather Station to measure atmospheric...

  6. Measurement of magnetic properties at cryogenic temperatures

    CERN Multimedia

    1977-01-01

    This picture shows part of the low-mu permeameter to measure permeability of stainless steels and other low-mu materials used in superconducting magnets. The sample, a 5 mm diam., 45 mm long rod, is suspended to long leads before being inserted in the test cryostat. For the measurement the sample is surrounded by a flux- measuring coil and placed in the field of a superconducting solenoid. At a given field the sample is removed.During the removal, the voltage induced in the flux-measuring coil is time integrated giving the flux variation. This equipment was developed to select stainless steels and other low-mu materials used in the ISR Prototype Superconducting Qaudrupole. The person is W.Ansorge.

  7. Identification of complex model thermal boundary conditions based on exterior temperature measurement

    International Nuclear Information System (INIS)

    Lu Jianming; Ouyang Guangyao; Zhang Ping; Rong Bojun

    2012-01-01

    Combining the advantages of the finite element software in temperature field analyzing with the multivariate function optimization arithmetic, a feasibility method based on the exterior temperature was proposed to get the thermal boundary conditions, which was required in temperature field analyzing. The thermal boundary conditions can be obtained only by some temperature measurement values. Taking the identification of the convection heat transfer coefficient of a high power density diesel engine cylinder head as an example, the calculation result shows that when the temperature measurement error was less than 0.5℃, the maximum relative error was less than 2%. It is shown that the new method was feasible (authors)

  8. Measuring the Densities of Aqueous Glasses at Cryogenic Temperatures.

    Science.gov (United States)

    Shen, Chen; Julius, Ethan F; Tyree, Timothy J; Dan, Ritwik; Moreau, David W; Thorne, Robert

    2017-06-28

    We demonstrate a method for determining the vitreous phase cryogenic temperature densities of aqueous mixtures, and other samples that require rapid cooling, to prepare the desired cryogenic temperature phase. Microliter to picoliter size drops are cooled by projection into a liquid nitrogen-argon (N2-Ar) mixture. The cryogenic temperature phase of the drop is evaluated using a visual assay that correlates with X-ray diffraction measurements. The density of the liquid N2-Ar mixture is adjusted by adding N2 or Ar until the drop becomes neutrally buoyant. The density of this mixture and thus of the drop is determined using a test mass and Archimedes principle. With appropriate care in drop preparation, management of gas above the liquid cryogen mixture to minimize icing, and regular mixing of the cryogenic mixture to prevent density stratification and phase separation, densities accurate to <0.5% of drops as small as 50 pL can readily be determined. Measurements on aqueous cryoprotectant mixtures provide insight into cryoprotectant action, and provide quantitative data to facilitate thermal contraction matching in biological cryopreservation.

  9. Measurement of temperature fluctuations and anomalous transport ...

    Indian Academy of Sciences (India)

    (b) Displacement of plasma from the centre of the vacuum vessel; horizontal displacement ¡А ( ve means ... vacuum vessel) and the ion saturation current Б× drawn by the pair is obtained by measur- ing the voltage drop ... sheared E- xB rotation (plasma rotation) as observed in other machines [8]. The fluctuation induced ...

  10. Upgrade of the cooling water temperature measures system for HLS

    International Nuclear Information System (INIS)

    Guo Weiqun; Liu Gongfa; Bao Xun; Jiang Siyuan; Li Weimin; He Duohui

    2007-01-01

    The cooling water temperature measures system for HLS (Hefei Light Source) adopts EPICS to the developing platform and takes the intelligence temperature cruise instrument for the front control instrument. Data of temperatures are required by IOCs through Serial Port Communication, archived and searched by Channel Archiver. The system can monitor the real-time temperatures of many channels cooling water and has the function of history data storage, and data network search. (authors)

  11. Spatial interpolation methods for monthly rainfalls and temperatures in Basilicata

    Directory of Open Access Journals (Sweden)

    Ferrara A

    2008-12-01

    Full Text Available Spatial interpolated climatic data on grids are important as input in forest modeling because climate spatial variability has a direct effect on productivity and forest growth. Maps of climatic variables can be obtained by different interpolation methods depending on data quality (number of station, spatial distribution, missed data etc. and topographic and climatic features of study area. In this paper four methods are compared to interpolate monthly rainfall at regional scale: 1 inverse distance weighting (IDW; 2 regularized spline with tension (RST; 3 ordinary kriging (OK; 4 universal kriging (UK. Besides, an approach to generate monthly surfaces of temperatures over regions of complex terrain and with limited number of stations is presented. Daily data were gathered from 1976 to 2006 period and then gaps in the time series were filled in order to obtain monthly mean temperatures and cumulative precipitation. Basic statistics of monthly dataset and analysis of relationship of temperature and precipitation to elevation were performed. A linear relationship was found between temperature and altitude, while no relationship was found between rainfall and elevation. Precipitations were then interpolated without taking into account elevation. Based on root mean squared error for each month the best method was ranked. Results showed that universal kriging (UK is the best method in spatial interpolation of rainfall in study area. Then cross validation was used to compare prediction performance of tree different variogram model (circular, spherical, exponential using UK algorithm in order to produce final maps of monthly precipitations. Before interpolating temperatures were referred to see level using the calculated lapse rate and a digital elevation model (DEM. The result of interpolation with RST was then set to originally elevation with an inverse procedure. To evaluate the quality of interpolated surfaces a comparison between interpolated and

  12. Development of electron temperature measuring system by silicon drift detector

    International Nuclear Information System (INIS)

    Song Xianying; Yang Jinwei; Liao Min

    2007-12-01

    Soft X-ray spectroscopy with two channels Silicon Drift Detector (SDD) are adopted for electron temperature measuring on HL-2A tokamak in 2005. The working principle, design and first operation of the SDD soft X-ray spectroscopy are introduced. The measuring results of electron temperature are also presented. The results show that the SDD is very good detector for electron temperature measuring on HL-2A tokamak. These will become a solid basic work to establish SDD array for electron temperature profiling. (authors)

  13. Measurement of relative permittivity of LTCC ceramic at different temperatures

    Science.gov (United States)

    Tan, Qiulin; Kang, Hao; Qin, Li; Xiong, Jijun; Zhou, Zhaoying; Zhang, Wendong; Luo, Tao; Xue, Chenyang; Liu, Jun

    2014-03-01

    Devices based on LTCC (low-temperature co-fired ceramic) technology are more widely applied in high temperature environments, and the temperature-dependent properties of the LTCC material play an important role in measurements of the characteristics of these devices at high temperature. In this paper, the temperature-dependence of the relative permittivity of DuPont 951 LTCC ceramic is studied from room temperature to 500 °C. An expression for relative permittivity is obtained, which relates the relative permittivity to the resonant frequency, inductance, parasitic capacitance and electrode capacitance of the LTCC sample. Of these properties, the electrode capacitance is the most strongly temperature-dependent. The LTCC sample resonant frequency, inductance and parasitic capacitance were measured (from room temperature to 500 °C) with a high temperature measurement system comprising a muffle furnace and network analyzer. We found that the resonant frequency reduced and the inductance and parasitic capacitance increased slightly as the temperature increases. The relative permittivity can be calculated from experimental frequency, inductance and parasitic capacitance measurements. Calculating results show that the relative permittivity of DuPont 951 LTCC ceramic ceramic increases to 8.21 from room temperature to 500 °C.

  14. Measurement of relative permittivity of LTCC ceramic at different temperatures

    Directory of Open Access Journals (Sweden)

    Qiulin Tan

    2014-02-01

    Full Text Available Devices based on LTCC (low-temperature co-fired ceramic technology are more widely applied in high temperature environments, and the temperature-dependent properties of the LTCC material play an important role in measurements of the characteristics of these devices at high temperature. In this paper, the temperature-dependence of the relative permittivity of DuPont 951 LTCC ceramic is studied from room temperature to 500 °C. An expression for relative permittivity is obtained, which relates the relative permittivity to the resonant frequency, inductance, parasitic capacitance and electrode capacitance of the LTCC sample. Of these properties, the electrode capacitance is the most strongly temperature-dependent. The LTCC sample resonant frequency, inductance and parasitic capacitance were measured (from room temperature to 500 °C with a high temperature measurement system comprising a muffle furnace and network analyzer. We found that the resonant frequency reduced and the inductance and parasitic capacitance increased slightly as the temperature increases. The relative permittivity can be calculated from experimental frequency, inductance and parasitic capacitance measurements. Calculating results show that the relative permittivity of DuPont 951 LTCC ceramic ceramic increases to 8.21 from room temperature to 500 °C.

  15. Measuring transient high temperature thermal phenomena in hostile environment

    International Nuclear Information System (INIS)

    Brenden, B.B.; Hartman, J.S.; Reich, F.R.

    1980-01-01

    The design of equipment for measuring temperature and strain in a rapidly heated and pressurized cylinder of stainless steel is discussed. Simultaneous cinematography of the full circumference of the cylinder without interference with temperature and strain measurements is also illustrated. The integrated system uses a reflective chamber for the sample and requires careful consideration of the spectral energy distribution utilized by each instrument

  16. Diagnostic accuracy of routine postoperative body temperature measurements

    NARCIS (Netherlands)

    Vermeulen, Hester; Storm-Versloot, Marja N.; Goossens, Astrid; Speelman, Peter; Legemate, Dink A.

    2005-01-01

    BACKGROUND: On surgical wards, body temperature is routinely measured, but there is no proof that this is useful for detecting postoperative infection. The aim of this study was to compare temperature measurements (the test) with the confirmed absence or presence of a postoperative infection (the

  17. An array for measuring detailed soil temperature profiles

    Science.gov (United States)

    Soil temperature dynamics can provide insights into soil variables which are much more difficult or impossible to measure. We designed an array to measure temperature at precise depth increments. Data was collected to determine if the construction materials influence surface and near-surface tempera...

  18. Thermocouple design for measuring temperatures of small insects

    Science.gov (United States)

    A.A. Hanson; R.C. Venette

    2013-01-01

    Contact thermocouples often are used to measure surface body temperature changes of insects during cold exposure. However, small temperature changes of minute insects can be difficult to detect, particularly during the measurement of supercooling points. We developed two thermocouple designs, which use 0.51 mm diameter or 0.127 mm diameter copper-constantan wires, to...

  19. Three-wavelength pyrometer for measuring flame temperatures.

    Science.gov (United States)

    Cashdollar, K L

    1979-08-01

    This paper describes a pyrometer that measures the continuum radiation from particles in a flame or explosion at three wavelengths (0.8 microm, 0.9 microm, and 1.0 microm). The particle temperature is calculated from the radiation data using the Planck equation. Temperatures measured for coal dust explosions in a closed vessel are presented.

  20. Low level TOC measurement method

    Science.gov (United States)

    Ekechukwu, Amy A.

    2001-01-01

    A method for the determination of total organic carbon in an aqueous sample by trapping the organic matter on a sorbent which is carbon free and analyzing the sorbent by combustion and determination of total CO.sub.2 by IR.

  1. Multi-channel temperature measurement system for automotive battery stack

    Science.gov (United States)

    Lewczuk, Radoslaw; Wojtkowski, Wojciech

    2017-08-01

    A multi-channel temperature measurement system for monitoring of automotive battery stack is presented in the paper. The presented system is a complete battery temperature measuring system for hybrid / electric vehicles that incorporates multi-channel temperature measurements with digital temperature sensors communicating through 1-Wire buses, individual 1-Wire bus for each sensor for parallel computing (parallel measurements instead of sequential), FPGA device which collects data from sensors and translates it for CAN bus frames. CAN bus is incorporated for communication with car Battery Management System and uses additional CAN bus controller which communicates with FPGA device through SPI bus. The described system can parallel measure up to 12 temperatures but can be easily extended in the future in case of additional needs. The structure of the system as well as particular devices are described in the paper. Selected results of experimental investigations which show proper operation of the system are presented as well.

  2. Assessment for Melting Temperature Measurement of Nucleic Acid by HRM

    Directory of Open Access Journals (Sweden)

    Jing Wang

    2016-01-01

    Full Text Available High resolution melting (HRM, with a high sensitivity to distinguish the nucleic acid species with small variations, has been widely applied in the mutation scanning, methylation analysis, and genotyping. For the aim of extending HRM for the evaluation of thermal stability of nucleic acid secondary structures on sequence dependence, we investigated effects of the dye of EvaGreen, metal ions, and impurities (such as dNTPs on melting temperature (Tm measurement by HRM. The accuracy of HRM was assessed as compared with UV melting method, and little difference between the two methods was found when the DNA Tm was higher than 40°C. Both insufficiency and excessiveness of EvaGreen were found to give rise to a little bit higher Tm, showing that the proportion of dye should be considered for precise Tm measurement of nucleic acids. Finally, HRM method was also successfully used to measure Tms of DNA triplex, hairpin, and RNA duplex. In conclusion, HRM can be applied in the evaluation of thermal stability of nucleic acid (DNA or RNA or secondary structural elements (even when dNTPs are present.

  3. Temperature sensitive surfaces and methods of making same

    Science.gov (United States)

    Liang, Liang [Richland, WA; Rieke, Peter C [Pasco, WA; Alford, Kentin L [Pasco, WA

    2002-09-10

    Poly-n-isopropylacrylamide surface coatings demonstrate the useful property of being able to switch charateristics depending upon temperature. More specifically, these coatings switch from being hydrophilic at low temperature to hydrophobic at high temperature. Research has been conducted for many years to better characterize and control the properties of temperature sensitive coatings. The present invention provides novel temperature sensitive coatings on articles and novel methods of making temperature sensitive coatings that are disposed on the surfaces of various articles. These novel coatings contain the reaction products of n-isopropylacrylamide and are characterized by their properties such as advancing contact angles. Numerous other characteristics such as coating thickness, surface roughness, and hydrophilic-to-hydrophobic transition temperatures are also described. The present invention includes articles having temperature-sensitve coatings with improved properties as well as improved methods for forming temperature sensitive coatings.

  4. Quantification of in situ temperature measurements on a PBI-based high temperature PEMFC unit cell

    DEFF Research Database (Denmark)

    Lebæk, Jesper; Ali, Syed Talat; Møller, Per

    2010-01-01

    The temperature is a very important operating parameter for all types of fuel cells. In the present work distributed in situ temperature measurements are presented on a polybenzimidazole based high temperature PEM fuel cell (HT-PEM). A total of 16 T-type thermocouples were embedded on both the an...

  5. Non-contact local temperature measurement inside an object using an infrared point detector

    Science.gov (United States)

    Hisaka, Masaki

    2017-04-01

    Local temperature measurement in deep areas of objects is an important technique in biomedical measurement. We have investigated a non-contact method for measuring temperature inside an object using a point detector for infrared (IR) light. An IR point detector with a pinhole was constructed and the radiant IR light emitted from the local interior of the object is photodetected only at the position of pinhole located in imaging relation. We measured the thermal structure of the filament inside the miniature bulb using the IR point detector, and investigated the temperature dependence at approximately human body temperature using a glass plate positioned in front of the heat source.

  6. Thermal conductivity measurement of liquid uranium dioxide by transient method

    International Nuclear Information System (INIS)

    Degiovanni, A.; Remy, B.

    2006-01-01

    This work deals with a new measurement method of the thermal conductivity of uranium dioxide in liquid phase. The sample, initially in the solid form, is heated above the melting point by a laser pulse. The temperature variation of the heated zone is measured with a fast pyrometer and allows to recover the thermal conductivity of the liquid using an inverse method. The uncertainty obtained by this method is significantly lower to the one encountered in the literature. (J.S.)

  7. Direct measurement high resolution wide range extreme temperature optical sensor using an all-silicon carbide probe.

    Science.gov (United States)

    Sheikh, Mumtaz; Riza, Nabeel A

    2009-05-01

    We propose and demonstrate a temperature sensing method using an all-silicon carbide probe that combines wavelength-tuned signal processing for coarse measurements and classical Fabry-Perot etalon peak shift for fine measurements. This method gives direct unambiguous temperature measurements with a high temperature resolution over a wide temperature range. Specifically, temperature measurements from room temperature to 1000 degrees C are experimentally demonstrated with an estimated resolution varying from 0.66 degrees C at room temperature to 0.12 degrees C at 1000 degrees C. The proposed sensor has applications in next-generation greener gas turbines for power production.

  8. Determination of soil evaporation fluxes using distributed temperature sensing methods

    Science.gov (United States)

    Serna, J. L.; Cristi Matte, F.; Munoz, J. F.; Suarez, F. I.

    2014-12-01

    The dynamics of evaporation fluxes in arid soils is an unresolved complex phenomenon that has a major impact on the basin's water availability. In arid zones, evaporation controls moisture contents near the soil surface and drives liquid water and water vapor fluxes through the vadose zone, playing a critical role in both the hydrological cycle and energy balance. However, determining soil evaporation in arid zones is a difficult undertaking. Thus, it is important to develop new measuring techniques that can determine evaporation fluxes. In the last decade, distributed temperature sensing (DTS) methods have been successfully used to investigate a wide range of hydrologic applications. In particular, DTS methods have been used indirectly to monitor soil moisture. Two methods have been developed: the passive and the active method. In the active mode, the DTS system uses cables with metal elements and a voltage difference is applied at the two ends to of the cable to heat it up for a defined time-period. Then, the cumulative temperature increase along the cable is computed and soil moisture is determined by using an empirical relation. DTS technology has also been used to determine water fluxes in porous media, but so far no efforts have been made to determine evaporation fluxes. Here, we investigate the feasibility of using the active DTS method to determine soil evaporation fluxes. To achieve this objective, column experiments were designed to study evaporation from sandy soils with shallow water tables. The soil columns were instrumented with traditional temperature and time-domain-reflectometry probes, and an armored fiber-optic cable that allows using the active method to estimate the soil moisture profile. In the experiments, the water table can be fixed at different depths and soil evaporation can be estimated by measuring the water added to the constant-head reservoir that feeds the column. Thus, allowing the investigation of soil evaporation fluxes from DTS

  9. Measuring temperatures with modified Kleiber 270B pyrometer

    International Nuclear Information System (INIS)

    Osch, E.V. van.

    1995-05-01

    At ECN a fast pyrometer is being used as a diagnostic tool for plasma disruption simulation experiments on candidate plasma facing materials for future thermonuclear fusion devices such as NET or ITER. The pyrometer is being used to measure the surface temperature response of the materials to short pulse high heat loads as induced by high power laser or electron beam, simulating the disrupting plasma's energy deposition. A procedure to measure surface temperatures without having to know surface emissivity in advance is described. The formulae needed in this procedure to obtain the correct temperature, starting from the initial incorrect temperature reading, are derived. Inversely, the formula to determine the emissivity of the surface when its temperature is known is equally derived. Finally, a small study on background level sensitivity is presented, showing the, in general, small effect of background on the temperature measurement. (orig.)

  10. Methods for measurement of durability parameters

    DEFF Research Database (Denmark)

    Hansen, Ernst Jan De Place

    1996-01-01

    Present selected methods for measurement of durabilty parameters relating to chlorides, corrosion, moisture and freeze-thaw, primarly on concrete. Advantages and drawbacks of the different methods are included.......Present selected methods for measurement of durabilty parameters relating to chlorides, corrosion, moisture and freeze-thaw, primarly on concrete. Advantages and drawbacks of the different methods are included....

  11. Measured gas and particle temperatures in VTT's entrained flow reactor

    DEFF Research Database (Denmark)

    Clausen, Sønnik; Sørensen, L.H.

    2006-01-01

    Particle and gas temperature measurements were carried out in experiments on VTTs entrained flow reactor with 5% and 10% oxygen using Fourier transform infrared emission spectroscopy (FTIR). Particle temperature measurements were performed on polish coal,bark, wood, straw particles, and bark...... and wood particles treated with additive. A two-color technique with subtraction of the background light was used to estimate particle temperatures during experiments. A transmission-emission technique was used tomeasure the gas temperature in the reactor tube. Gas temperature measurements were in good...... agreement with thermocouple readings. Gas lines and bands from CO, CO2 and H2O can be observed in the spectra. CO was only observed at the first measuring port (100ms) with the strongest CO-signal seen during experiments with straw particles. Variations in gas concentration (CO2 and H2O) and the signal from...

  12. High temperature cooling system and method

    Science.gov (United States)

    Loewen, Eric P.

    2006-12-12

    A method for cooling a heat source, a method for preventing chemical interaction between a vessel and a cooling composition therein, and a cooling system. The method for cooling employs a containment vessel with an oxidizable interior wall. The interior wall is oxidized to form an oxide barrier layer thereon, the cooling composition is monitored for excess oxidizing agent, and a reducing agent is provided to eliminate excess oxidation. The method for preventing chemical interaction between a vessel and a cooling composition involves introducing a sufficient quantity of a reactant which is reactive with the vessel in order to produce a barrier layer therein that is non-reactive with the cooling composition. The cooling system includes a containment vessel with oxidizing agent and reducing agent delivery conveyances and a monitor of oxidation and reduction states so that proper maintenance of a vessel wall oxidation layer occurs.

  13. Feedwater temperature control methods and systems

    Science.gov (United States)

    Moen, Stephan Craig; Noonan, Jack Patrick; Saha, Pradip

    2014-04-22

    A system for controlling the power level of a natural circulation boiling water nuclear reactor (NCBWR) is disclosed. The system, in accordance with an example embodiment of the present invention, may include a controller configured to control a power output level of the NCBWR by controlling a heating subsystem to adjust a temperature of feedwater flowing into an annulus of the NCBWR. The heating subsystem may include a steam diversion line configured to receive steam generated by a core of the NCBWR and a steam bypass valve configured to receive commands from the controller to control a flow of the steam in the steam diversion line, wherein the steam received by the steam diversion line has not passed through a turbine. Additional embodiments of the invention may include a feedwater bypass valve for controlling an amount of flow of the feedwater through a heater bypass line to the annulus.

  14. A Simple and Accurate Method for Measuring Enzyme Activity.

    Science.gov (United States)

    Yip, Din-Yan

    1997-01-01

    Presents methods commonly used for investigating enzyme activity using catalase and presents a new method for measuring catalase activity that is more reliable and accurate. Provides results that are readily reproduced and quantified. Can also be used for investigations of enzyme properties such as the effects of temperature, pH, inhibitors,…

  15. Heat-capacity measurements on small samples: The hybrid method

    NARCIS (Netherlands)

    Klaasse, J.C.P.; Brück, E.H.

    2008-01-01

    A newly developed method is presented for measuring heat capacities on small samples, particularly where thermal isolation is not sufficient for the use of the traditional semiadiabatic heat-pulse technique. This "hybrid technique" is a modification of this heat-pulse method in case the temperature

  16. Noncontact measurement of high temperature using optical fiber sensors

    Science.gov (United States)

    Claus, R. O.

    1990-01-01

    The primary goal of this research program was the investigation and application of noncontact temperature measurement techniques using optical techniques and optical fiber methods. In particular, a pyrometer utilizing an infrared optical light pipe and a multiwavelength filtering approach was designed, revised, and tested. This work was motivated by the need to measure the temperatures of small metallic pellets (approximately 3 mm diameter) in free fall at the Microgravity Materials Processing Drop Tube at NASA Marshall Space Flight Center. In addition, research under this program investigated the adaptation of holography technology to optical fiber sensors, and also examined the use of rare-earth dopants in optical fibers for use in measuring temperature. The pyrometer development effort involved both theoretical analysis and experimental tests. For the analysis, a mathematical model based on radiative transfer principles was derived. Key parameter values representative of the drop tube system, such as particle size, tube diameter and length, and particle temperature, were used to determine an estimate of the radiant flux that will be incident on the face of an optical fiber or light pipe used to collect radiation from the incandescent falling particle. An extension of this work examined the advantage of inclining or tilting the collecting fiber to increase the time that the falling particle remains in the fiber field-of-view. Those results indicate that increases in total power collected of about 15 percent may be realized by tilting the fiber. In order to determine the suitability of alternative light pipes and optical fibers, and experimental set-up for measuring the transmittance and insertion loss of infrared fibers considered for use in the pyrometer was assembled. A zirconium fluoride optical fiber and several bundles of hollow core fiber of varying diameters were tested. A prototype two-color pyrometer was assembled and tested at Virginia Tech, and then

  17. An interlaboratory comparison of methods for measuring rock matrix porosity

    International Nuclear Information System (INIS)

    Rasilainen, K.; Hellmuth, K.H.; Kivekaes, L.; Ruskeeniemi, T.; Melamed, A.; Siitari-Kauppi, M.

    1996-09-01

    An interlaboratory comparison study was conducted for the available Finnish methods of rock matrix porosity measurements. The aim was first to compare different experimental methods for future applications, and second to obtain quality assured data for the needs of matrix diffusion modelling. Three different versions of water immersion techniques, a tracer elution method, a helium gas through-diffusion method, and a C-14-PMMA method were tested. All methods selected for this study were established experimental tools in the respective laboratories, and they had already been individually tested. Rock samples for the study were obtained from a homogeneous granitic drill core section from the natural analogue site at Palmottu. The drill core section was cut into slabs that were expected to be practically identical. The subsamples were then circulated between the different laboratories using a round robin approach. The circulation was possible because all methods were non-destructive, except the C-14-PMMA method, which was always the last method to be applied. The possible effect of drying temperature on the measured porosity was also preliminarily tested. These measurements were done in the order of increasing drying temperature. Based on the study, it can be concluded that all methods are comparable in their accuracy. The selection of methods for future applications can therefore be based on practical considerations. Drying temperature seemed to have very little effect on the measured porosity, but a more detailed study is needed for definite conclusions. (author) (4 refs.)

  18. A Model-Based Temperature-Prediction Method by Temperature-Induced Spectral Variation and Correction of the Temperature Effect.

    Science.gov (United States)

    Yang, Qhi-xiao; Peng, Si-long; Shan, Peng; Bi, Yi-ming; Tang, Liang; Xie, Qiong

    2015-05-01

    In the present paper, a new model-based method was proposed for temperature prediction and correction. First, a temperature prediction model was obtained from training samples; then, the temperature of test samples were predicted; and finally, the correction model was used to reduce the nonlinear effects of spectra from temperature variations. Two experiments were used to verify the proposed method, including a water-ethanol mixture experiment and a ternary mixture experiment. The results show that, compared with classic method such as continuous piecewise direct standardization (CPDS), our method is efficient for temperature correction. Furthermore, the temperatures of test samples are not necessary in the proposed method, making it easier to use in real applications.

  19. Noncontact true temperature measurement. [of levitated sample using laser pyrometer

    Science.gov (United States)

    Lee, Mark C.; Allen, James L.

    1987-01-01

    A laser pyrometer has been developed for acquiring the true temperature of a levitated sample. The laser beam is first expanded to cover the entire cross-sectional surface of the target. For calibration of such a system, the reflectivity signal of an ideal 0.95 cm diameter gold-coated sphere (reflectivity = 0.99) is used as the reference for any other real targets. The emissivity of the real target can then be calculated. The overall system constant is obtained by passively measuring the radiance of a blackbody furnace (emissivity = 1.0) at a known, arbitrary temperature. Since the photo sensor used is highly linear over the entire operating temperature range, the true temperature of the target can then be computed. Preliminary results indicate that true temperatures thus obtained are in excellent correlation with thermocouple measured temperatures.

  20. Applicable technical method for lower temperature freeze ...

    African Journals Online (AJOL)

    Cryo-fixation and freeze substitution followed by microscopy are commonly used sample preparation methods for visualizing the morphology of intracellular organelles. Freeze substitution is an especially important preparative step because it enables the preservation of intracellular structures in cryo-fixed cells close to the ...

  1. Measuring electric conductivity in liquid metals by eddy current method

    International Nuclear Information System (INIS)

    Zhuravlev, S.P.; Ostrovskij, O.I.; Grigoryan, V.A.

    1982-01-01

    Technique permitting to apply the method of vertiginous currents for investigation of electric conductivity of metal melts in the high temperature range is presented. Interferences affecting accuracy of measurements are specified and ways of their removing are pointed out. Scheme of measuring and design of the facility are described. Results of measuring electric resistance of liquid Fe, Co, Ni obtained for the first time by this method are presented. The data obtained agree with the results of measurements conducted by the method of the rotating magnetic field. Difference in absolute values of electric resistance in parallel experiments for each metal does not exceed 4%

  2. A System to Measure Both Inner and Outer Car Tire Temperatures ``in situ''

    Science.gov (United States)

    Koštial, P.; Mokryšová, M.; Šišáková, J.; Mošková, Z.; Rusnáková, S.

    2009-02-01

    In the paper, a system for the complex analysis of the internal and external tire temperatures and pressure of sporty tires is presented. Tests were performed on the test circuit of a tire producer. The CTPA 05 measuring system (complex temperature-pressure analyzer) enables simultaneous measurements of the internal temperature and pressure in a passenger or sports tire. The experimentalist determines that the CTPA 05 can be used to measure independently the external temperature of the overcoat on the front wheel driving tires at three points. Measurements of both the internal tire temperature and pressure, as well as of the external tire temperature, are collected together with GPS (global position system) data. The system of measurement is fully automatic and contactless. The obtained results are in very good agreement with those obtained by independent methods.

  3. Tooth Whitening And Temperature Rise With Two Bleaching Activation Methods

    Science.gov (United States)

    Abu-ElMagd, D. M.; El-Sayad, I. I.; Abd El-Gawad, L. M.

    2009-09-01

    To measure the tooth whitening and the surface and Intrapulpal temperature increase in vitro on freshly extracted upper human central incisors after chemical, Zoom AP light and diode laser activated bleaching. Thirty caries-free upper human incisors were selected. Teeth were divided into three equal groups according to the methods of activation of the bleaching agent (n = 10). A whitening gel containing hydrogen peroxide was applied to the buccal surface of all teeth. Group I was bleached using chemically activated hydrogen peroxide gel, for three applications of 15 min each. Group II was bleached with high intensity advanced power Zoom activation light (Zoom AP), for three applications of 15 min each. Group III was bleached with diode laser activation technique, where the teeth were irradiated with 2 Watt diode laser for three applications of 30 sec each. The whitening degree was assessed using an image analysis system, while temperature rise was recorded using a thermocouple on the external tooth surface and Intrapulpal. The degree of whitening increased significantly in all groups. However, the percentage of whitening was not statistically significantly different between the three groups. In addition, group II showed statistically significant higher mean rise in both surface and pulp temperatures than group I and group III. Chemical bleaching produces the same whitening effect as Zoom AP light and laser, with no surface or pulpal temperature rise. Laser application is faster and produces less surface and pulp temperature increase than Zoom AP light. Diode laser used to activate bleaching gels is not considered dangerous to the vitality of dental pulp using power settings of 2 W.

  4. Surface temperature measurements of heterogeneous explosives by IR emission

    Energy Technology Data Exchange (ETDEWEB)

    Henson, B.F.; Funk, D.J.; Dickson, P.M.; Fugard, C.S.; Asay, B.W.

    1998-03-01

    The authors present measurements of the integrated IR emission (1--5 {micro}m) from both the heterogeneous explosive PBX 9501 and pure HMX at calibrated temperatures from 300 C to 2,500 C. The IR power emitted as a function of temperature is that expected of a black body, attenuated by a unique temperature independent constant which the authors report as the thermal emissivity. The authors have utilized this calibration of IR emission in measurements of the surface temperature from PBX 9501 subject to 1 GPa, two dimensional impact, and spontaneous ignition in unconfined cookoff. They demonstrate that the measurement of IR emission in this spectral region provides a temperature probe of sufficient sensitivity to resolve the thermal response from the solid explosive throughout the range of weak mechanical perturbation, prolonged heating to ignition, and combustion.

  5. The electronic temperature control and measurements reactor fuel rig circuits

    International Nuclear Information System (INIS)

    Glowacki, S.W.

    1980-01-01

    The electronic circuits of two digital temperature meters developed for the thermocouple of Ni-NiCr type are described. The output thermocouple signal as converted by means of voltage-to-freguency converter. The frequency is measured by a digital scaler controled by quartz generator signals. One of the described meter is coupled with digital temperature controler which drives the power stage of the reactor rig heater. The internal rig temperature is measured by the thermocouple providing the input signal to the mentioned voltage-to-frequency converter, that means the circuits work in the negative feedback loop. The converter frequency-to-voltage ratio is automatically adjusted to match to thermocouple sensitivity changes in the course of the temperature variations. The accuracy of measuring system is of order of +- 1degC for thermocouple temperature changes from 523 K up to 973 K (50degC up to 700degC). (author)

  6. Acoustic temperature profile measurement technique for large combustion chambers

    Science.gov (United States)

    Venkateshan, S. P.; Shakkottai, P.; Kwack, E. Y.; Back, L. H.

    1989-01-01

    Measurement of times of flight of sound waves can be used to determine temperatures in a gas. This paper describes a system, based on this principle, that is capable of giving the temperature profile in a nonisothermal gas volume, for example, prevalent in a large furnace. The apparatus is simple, rugged, accurate, and capable of being automated for process control applications. It is basically an acoustic waveguide where the outside temperature profile is transferred to a chosen gas contained inside the guide.

  7. Selective solar absorber emittance measurement at elevated temperature

    Science.gov (United States)

    Giraud, Philémon; Braillon, Julien; Raccurt, Olivier

    2017-06-01

    Durability of solar components for CSP (Concentrated Solar Power Plant) technologies is a key point to lower cost and ensure their large deployment. These technologies concentrated the solar radiation by means of mirrors on a receiver tube where it is collected as thermal energy. The absorbers are submitted to strong environmental constraints and the degradation of their optical properties (emittance and solar absorbance) have a direct impact on performance. The characterization of a material in such condition is complicated and requires advanced apparatuses, and different measurement methods exist for the determination of the two quantities of relevance regarding an absorber, which are its emittance and its solar absorbance. The objective is to develop new optical equipment for measure the emittance of this solar absorber at elevated temperature. In this paper, we present an optical bench developed for emittance measurement on absorbers is conditions of use. Results will be shown, with a discussion of some factors of influence over this measurement and how to control them.

  8. Technology and education: First approach for measuring temperature with Arduino

    Science.gov (United States)

    Carrillo, Alejandro

    2017-04-01

    This poster session presents some ideas and approaches to understand concepts of thermal equilibrium, temperature and heat in order to bulid a man-nature relationship in a harmonious and responsible manner, emphasizing the interaction between science and technology, without neglecting the relationship of the environment and society, an approach to sustainability. It is proposed the development of practices that involve the use of modern technology, of easy access and low cost to measure temperature. We believe that the Arduino microcontroller and some temperature sensors can open the doors of innovation to carry out such practices. In this work we present some results of simple practices presented to a population of students between the ages of 16 and 17 years old. The practices in this proposal are: Zero law of thermodynamics and the concept of temperature, calibration of thermometers and measurement of temperature for heating and cooling of three different substances under the same physical conditions. Finally the student is asked to make an application that involves measuring of temperature and other physical parameters. Some suggestions are: to determine the temperature at which we take some food, measure the temperature difference at different rooms of a house, housing constructions that favour optimal condition, measure the temperature of different regions, measure of temperature trough different colour filters, solar activity and UV, propose applications to understand current problems such as global warming, etc. It is concluded that the Arduino practices and electrical sensors increase the cultural horizon of the students while awaking their interest to understand their operation, basic physics and its application from a modern perspective.

  9. Apparatus and test method for characterizing the temperature regulating properties of thermal functional porous polymeric materials.

    Science.gov (United States)

    Yao, Bao-Guo; Zhang, Shan; Zhang, De-Pin

    2017-05-01

    In order to evaluate the temperature regulating properties of thermal functional porous polymeric materials such as fabrics treated with phase change material microcapsules, a new apparatus was developed. The apparatus and the test method can measure the heat flux, temperature, and displacement signals during the dynamic contact and then quickly give an evaluation for the temperature regulating properties by simulating the dynamic heat transfer and temperature regulating process when the materials contact the body skin. A series of indices including the psychosensory intensity, regulating capability index, and relative regulating index were defined to characterize the temperature regulating properties. The measurement principle, the evaluation criteria and grading method, the experimental setup and the test results discussion, and the gage capability analysis of the apparatus are presented. The new apparatus provides a method for the objective measurement and evaluation of the temperature regulating properties of thermal functional porous polymeric materials.

  10. Magnetometry and electrical transport measurements of high temperature superconductors

    Energy Technology Data Exchange (ETDEWEB)

    Chan, Mun Keat [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-02-13

    Prior to preparing and performing measurements in pulsed magnetic fields, it is necessary to characterize them. For the cuprates such as HgBa2CuO4+δ (Hg1201), measurements of the superconducting temperature is essential. This experiment comprises just such a characterization of Hg1201 crystals to be used for pulsed magnetic field measurements.

  11. Two particle correlation studies and measurement of temperature

    International Nuclear Information System (INIS)

    Rana, T.K.; Bhattacharya, C.; Kundu, S.; Banerjee, K.; Dey, A.; Ghosh, T.K.; Mukherjee, G.; Gupta, D.; Meena, J.K.; Roy, A.; Dhara, P.; Banerjee, S.R.; Bhattacharya, S.

    2007-01-01

    In order to obtain more detailed information about the emission temperatures, we have measured two particle correlation function at small relative momenta for the reaction 20 Ne+ 12 C at 145 MeV. Here, we report the temperature extracted from the correlation studies and compared it with those obtained from slope thermometer

  12. Miniature ingestible telemeter devices to measure deep-body temperature

    Science.gov (United States)

    Pope, J. M.; Fryer, T. B. (Inventor)

    1976-01-01

    A telemetry device comprised of a pill-size ingestible transmitter developed to obtain deep body temperature measurements of a human is described. The device has particular utility in the medical field where deep body temperatures provide an indication of general health.

  13. Measurement of local void fraction at elevated temperature and pressure

    International Nuclear Information System (INIS)

    Duncan, D.; Trabold, T.A.

    1993-03-01

    Significant advances have recently been made in analytical and computational methods for the prediction of local thermal-hydraulic conditions in gas/liquid two-phase flows. There is, however, a need for extensive experimental data, for the dual purposes of constitutive relation development and code qualification. There is especially true of systems involving complicated geometries and/or extreme flow conditions for which little, if any, applicable information exists in the open literature. For the tests described in the present paper, a novel electrical probe has been applied to measure the void fraction in atmospheric pressure air/water flows, and steam/water mixtures at high temperature and pressure. The data acquired in the latter experiments are compared with the results of a one-dimensional two-fluid computational analysis

  14. Time-Resolved Surface Temperature Measurement for Pulsed Ablative Thrusters

    National Research Council Canada - National Science Library

    Antonsen, Erik

    2003-01-01

    .... The diagnostic draws on heritage from the experimental dynamic crack propagation community which has used photovoltaic infrared detectors to measure temperature rise in materials in the process of fracture...

  15. On bias of kinetic temperature measurements in complex plasmas

    DEFF Research Database (Denmark)

    Kantor, M.; Moseev, D.; Salewski, Mirko

    2014-01-01

    The kinetic temperature in complex plasmas is often measured using particle tracking velocimetry. Here, we introduce a criterion which minimizes the probability of faulty tracking of particles with normally distributed random displacements in consecutive frames. Faulty particle tracking results i...

  16. Lower atmospheric temperature profile measurements using a Raman lidar

    Science.gov (United States)

    Melfi, S. H.; Whiteman, D.

    1986-01-01

    A Raman lidar system was used to measure the temperature profile of the upper troposphere and lower stratosphere. The system consists of a tripled Nd-YAG laser and a 1.5 meter diameter telescope. Two photomultipliers are used at the output of the telescope to allow for measurements at both the laser wavelength and at the Raman shifted wavelength due to atmospheric nitrogen. The signal from the photomultipliers is recorded as photon counts in 1 microsec bins. The results of a number of laser shots are summed together to provide atmospheric returns which have acceptable signal to noise characteristics. Measurements of the Raman nitrogen return were acquired up to an altitude in excess of 20 km. Temperature profiles were retrieved from the attenuation corrected Raman nitrogen return assuming the atmosphere to be in hydrostatic equilibrium and using the ideal gas law. Retrieved temperature profiles are shown compared with independent temperature measurements.

  17. Combined NMR moisture, temperature and pressure measurements during heating

    Directory of Open Access Journals (Sweden)

    Pel L.

    2013-09-01

    Full Text Available For model validation, quantitative measurements of the evolution of moisture, temperature, and pressure distributions in time are needed. For this purpose, we have developed an NMR setup to measure the moisture transport in heated building materials. The measured combined moisture content and temperature profiles give a unique insight in the moisture transport and dehydration kinetics inside concrete during fire. These measurements give the first quantitative proof for the build-up of a moisture peak due to the vapor pressure build-up. In this study we have also combined for the first time the measurement of the moisture and temperature profiles with the measurement of the pressure at one position, which show that the pressure build up is directly related to the moisture profiles.

  18. Calibration Method and Uncertainty Assessment of a High-Temperature GHP Apparatus

    Science.gov (United States)

    Yao, Kai; Zheng, Huibao; Liu, Yunchuan; Meng, Xiangyan; Zhou, Yanping; Wang, Xuerong; Wang, Kang; Wang, Qianqian

    2018-02-01

    In this research, a calibration method of a high-temperature guarded hot plate (GHP) apparatus was proposed in order to improve the measurement accuracy of thermal conductivity. The measurement uncertainties of this GHP apparatus were assessed to validate the reliability of this calibration method. The temperature difference across the guarded gap was set as the bias value to eliminate the heat exchange over the guarded gap. The effects of the thermal expansion and pressure of the apparatus on thickness were investigated to revise the measurement results of in- situ thickness and meter area, respectively. The assessed uncertainty indicated that the related expanded uncertainty approximately increased with the increase in testing temperature and the calibration method should be valid in the temperature range. The contribution of each factor on the combined uncertainty showed that the temperature distribution in plane direction was the main factor in influencing the measurement of thermal conductivity.

  19. A high-stability non-contact dilatometer for low-amplitude temperature-modulated measurements

    Energy Technology Data Exchange (ETDEWEB)

    Luckabauer, Martin; Sprengel, Wolfgang; Würschum, Roland [Institute of Materials Physics, Graz University of Technology, A-8010 Graz (Austria)

    2016-07-15

    Temperature modulated thermophysical measurements can deliver valuable insights into the phase transformation behavior of many different materials. While especially for non-metallic systems at low temperatures numerous powerful methods exist, no high-temperature device suitable for modulated measurements of bulk metallic alloy samples is available for routine use. In this work a dilatometer for temperature modulated isothermal and non-isothermal measurements in the temperature range from room temperature to 1300 K is presented. The length measuring system is based on a two-beam Michelson laser interferometer with an incremental resolution of 20 pm. The non-contact measurement principle allows for resolving sinusoidal length change signals with amplitudes in the sub-500 nm range and physically decouples the length measuring system from the temperature modulation and heating control. To demonstrate the low-amplitude capabilities, results for the thermal expansion of nickel for two different modulation frequencies are presented. These results prove that the novel method can be used to routinely resolve length-change signals of metallic samples with temperature amplitudes well below 1 K. This high resolution in combination with the non-contact measurement principle significantly extends the application range of modulated dilatometry towards high-stability phase transformation measurements on complex alloys.

  20. Temperature and voltage measurement in quantum systems far from equilibrium

    Science.gov (United States)

    Shastry, Abhay; Stafford, Charles A.

    2016-10-01

    We show that a local measurement of temperature and voltage for a quantum system in steady state, arbitrarily far from equilibrium, with arbitrary interactions within the system, is unique when it exists. This is interpreted as a consequence of the second law of thermodynamics. We further derive a necessary and sufficient condition for the existence of a solution. In this regard, we find that a positive temperature solution exists whenever there is no net population inversion. However, when there is a net population inversion, we may characterize the system with a unique negative temperature. Voltage and temperature measurements are treated on an equal footing: They are simultaneously measured in a noninvasive manner, via a weakly coupled thermoelectric probe, defined by requiring vanishing charge and heat dissipation into the probe. Our results strongly suggest that a local temperature measurement without a simultaneous local voltage measurement, or vice versa, is a misleading characterization of the state of a nonequilibrium quantum electron system. These results provide a firm mathematical foundation for voltage and temperature measurements far from equilibrium.

  1. Ring to measure magnetic permeability at cryogenic temperatures

    CERN Multimedia

    1977-01-01

    While for magn. permeability measurements at room temperature a split-coil permeameter is used (see photo 7708553X), for measurements at cryogenic temperatures the excitation and the flux-measuring coils are wound directly on the ring sample by means of a toroidal winding machine. The ring in the picture was made to select the mild steel for the ISR Prototype Superconducting Quadrupole(see photo 7702690X). The excitation coil was wound with 1 mm diam. copper wire and had about 2730 turns. For measurements at 4.2 K a max. current of 90 A was used. See also photos 7708553X,7708100,7708103.

  2. Diagnostic of the temperature and differential emission measure (DEM based on Hinode/XRT data

    Directory of Open Access Journals (Sweden)

    P. Rudawy

    2008-10-01

    Full Text Available We discuss here various methodologies and an optimal strategy of the temperature and emission measure diagnostics based on Hinode X-Ray Telescope data. As an example of our results we present the determination of the temperature distribution of the X-rays emitting plasma using a filters ratio method and three various methods of the calculation of the differential emission measure (DEM. We have found that all these methods give results similar to the two filters ratio method. Additionally, all methods of the DEM calculation gave similar solutions. We can state that the majority of the pairs of the Hinode filters allows one to derive the temperature and emission measure in the isothermal plasma approximation using standard diagnostics based on the two filters ratio method. In cases of strong flares one can also expect good conformity of the results obtained using a Withbroe – Sylwester, genetic algorithm and least-squares methods of the DEM evaluation.

  3. Analytical method for turbine blade temperature mapping to estimate a pyrometer input signal

    OpenAIRE

    MacKay, James D.

    1987-01-01

    The purpose of this thesis is to develop a method to estimate local blade temperatures in a gas turbine for comparison with the output signal of an experimental pyrometer. The goal of the method is to provide a temperature measurement benchmark based on a knowledge of blade geometry and engine operating conditions. A survey of currently available methods is discussed including both experimental and analytical techniques.The purpose of this thesis is to develop a method to estim...

  4. Method of measuring luminescence of a material

    Science.gov (United States)

    Miller, Steven D.

    2015-12-15

    A method of measuring luminescence of a material is disclosed. The method includes applying a light source to excite an exposed material. The method also includes amplifying an emission signal of the material. The method further includes measuring a luminescent emission at a fixed time window of about 10 picoseconds to about 10 nanoseconds. The luminescence may be radio photoluminescence (RPL) or optically stimulated luminescence (OSL).

  5. Noninvasive ultrasonic measurements of temperature distribution and heat fluxes in nuclear systems

    International Nuclear Information System (INIS)

    Jia, Yunlu; Skliar, Mikhail

    2015-01-01

    Measurements of temperature and heat fluxes through structural materials are important in many nuclear systems. One such example is dry storage casks (DSC) that are built to store highly radioactive materials, such as spent nuclear reactor fuel. The temperature inside casks must be maintained within allowable limits of the fuel assemblies and the DSC components because many degradation mechanisms are thermally controlled. In order to obtain direct, real-time measurements of temperature distribution without insertion of sensing elements into harsh environment of storage casks, we are developing noninvasive ultrasound (US) methods for measuring spatial distribution of temperature inside solid materials, such as concrete overpacks, steel casings, thimbles, and rods. The measured temperature distribution can then be used to obtain heat fluxes that provide calorimetric characterisation of the fuel decay, fuel distribution inside the cask, its integrity, and accounting of nuclear materials. The physical basis of the proposed approach is the temperature dependence of the speed of sound in solids. By measuring the time it takes an ultrasound signal to travel a known distance between a transducer and a receiver, the indication about the temperature distribution along the path of the ultrasound propagation may be obtained. However, when temperature along the path of US propagation is non-uniform, the overall time of flight of an ultrasound signal depends on the temperature distribution in a complex and unknown way. To overcome this difficulty, the central idea of our method is to create an US propagation path inside material of interest which incorporates partial ultrasound reflectors (back scatterers) at known locations and use the train of created multiple echoes to estimate the temperature distribution. In this paper, we discuss experimental validation of this approach, the achievable accuracy and spatial resolution of the measured temperature profile, and stress the

  6. FFTF fuel assembly outlet temperature measurements and comparison to predictions

    International Nuclear Information System (INIS)

    Cramer, E.R.

    1984-06-01

    The data from the FFTF core outlet thermocouples have been valuable in verifying the performance of the core assemblies. The data have been useful to the experimental program and as an aid in understanding some reactor operating phenomena. The thermocouple reliability and repeatability have been good. Almost all of the fueled positions in the core have 3 operable thermocouples and every position has at least one. Differences between the measured assembly outlet temperatures and the calculated outlet temperatures have generally been small. Where significant differences have occurred, an explanation has been found. The difference between the measured and calculated outlet temperatures for each assembly remains constant during the cycle

  7. Evaluation of Raytek infrared pyrometer for continuous propellant temperature measurement

    Science.gov (United States)

    Dykstra, Mark D.

    1990-01-01

    The primary purpose of this evaluation was to determine if the Raytek IR pyrometer that was installed in the 600 gallon propellant mixers could be used to provide a continuous, accurate, reliable measurement of the propellant temperature during mixing. The Raytek infrared sensor is not recommended to be used for controlling propellant temperature nor for inspection buy-off. The first part of the evaluation was to determine the accuracy of the sensor in measuring the propellant temperature. The second part was to determine the reliability of the air purge design in preventing contamination of the IR window.

  8. Measurement of Temperature Dependent Apparent Specific Heat Capacity in Electrosurgery.

    Science.gov (United States)

    Karaki, Wafaa; Akyildiz, Ali; Borca Tasciuc, Diana-Andra; De, Suvranu

    2016-01-01

    This paper reports on the measurement of temperature dependent apparent specific heat of ex-vivo porcine liver tissue during radiofrequency alternating current heating for a large temperature range. The difference between spatial and temporal evolution of experimental temperature, obtained during electrosurgical heating by infrared thermometry, and predictions based on finite element modeling was minimized to obtain the apparent specific heat. The model was based on transient heat transfer with internal heat generation considering heat storage along with conduction. Such measurements are important to develop computational models for real time simulation of electrosurgical procedures.

  9. Influence of temperature to quenching on liquid scintillation measurement

    CERN Document Server

    Kato, T

    2003-01-01

    The amount of quench is measured with liquid scintillation spectrometer changing the temperature of the sample. The range of the changed temperature is between 0 deg C and 35 deg C. The measurement is carried out for three kinds of unquenched standard, two quenched standards and fifteen kinds of scintillation cocktail and the mixed sample. It is confirmed that the amount of quench increases for all samples as the temperature rises. The influence of the changed amount of quench to the quench correction is examined. (author)

  10. Modeling the wet bulb globe temperature using standard meteorological measurements.

    Science.gov (United States)

    Liljegren, James C; Carhart, Richard A; Lawday, Philip; Tschopp, Stephen; Sharp, Robert

    2008-10-01

    The U.S. Army has a need for continuous, accurate estimates of the wet bulb globe temperature to protect soldiers and civilian workers from heat-related injuries, including those involved in the storage and destruction of aging chemical munitions at depots across the United States. At these depots, workers must don protective clothing that increases their risk of heat-related injury. Because of the difficulty in making continuous, accurate measurements of wet bulb globe temperature outdoors, the authors have developed a model of the wet bulb globe temperature that relies only on standard meteorological data available at each storage depot for input. The model is composed of separate submodels of the natural wet bulb and globe temperatures that are based on fundamental principles of heat and mass transfer, has no site-dependent parameters, and achieves an accuracy of better than 1 degree C based on comparisons with wet bulb globe temperature measurements at all depots.

  11. In situ gas temperature measurements by UV-absorption spectroscopy

    DEFF Research Database (Denmark)

    Fateev, Alexander; Clausen, Sønnik

    2009-01-01

    The absorption spectrum of the NO A(2)Sigma(+) <- X(2)Pi gamma-system can be used for in situ evaluation of gas temperature. Experiments were performed with a newly developed atmospheric-pressure high-temperature flow gas cell at highly uniform and stable gas temperatures over a 0.533 m path in t....... The accuracy of both methods is discussed. Validation of the classical Lambert-Beer law has been demonstrated at NO concentrations up to 500 ppm and gas temperatures up to 1,500 degrees C over an optical absorption path length of 0.533 m....

  12. Simultaneous measurement of gas concentration and temperature by the ball surface acoustic wave sensor

    Science.gov (United States)

    Yamanaka, Kazushi; Akao, Shingo; Takeda, Nobuo; Tsuji, Toshihiro; Oizumi, Toru; Tsukahara, Yusuke

    2017-07-01

    We have developed a ball surface acoustic wave (SAW) trace moisture sensor with an amorphous silica sensitive film and realized wide-range measurement from 0.017 ppmv [a frost point (FP) of -99 °C] to 6.0 × 103 ppmv (0 °C FP). However, since the sensitivity of the sensor depends on the temperature, measurement results are disturbed when the temperature largely changes. To overcome this problem, we developed a method to simultaneously measure temperature and gas concentration using a ball SAW sensor. Temperature and concentration is derived by solving equations for the delay time change at two frequencies. When the temperature had a large jump, the delay time change was significantly disturbed, but the water concentration was almost correctly measured, by compensating the sensitivity change using measured temperature. The temperature measured by a ball SAW sensor will also be used to control the ball temperature. This method will make a ball SAW sensor reliable in environments of varying temperatures.

  13. Low-temperature mobility measurements on CMOS devices

    International Nuclear Information System (INIS)

    Hairpetian, A.; Gitlin, D.; Viswanathan, C.R.

    1989-01-01

    The surface channel mobility of carriers in eta- and rho-MOS transistors fabricated in a CMOS process was accurately determined at low temperatures down to 5 Κ. The mobility was obtained by an accurate measurement of the inversion charge density using a split C-V technique and the conductance at low drain voltages. The split C-V technique was validated at all temperatures using a one-dimensional Poisson solver (MOSCAP), which was modified for low-temperature application. The mobility dependence on the perpendicular electric field for different substrate bias values appears to have different temperature dependence for eta- and rho-channel devices. The electron mobility increases with a decrease in temperature at all gate voltages. On the other hand, the hole mobility exhibits a different temperature behavior depending upon whether the gate voltage corresponds to strong inversion or is near threshold

  14. Non-invasive continuous core temperature measurement by zero heat flux

    NARCIS (Netherlands)

    Teunissen, L.P.J.; Klewer, J.; Haan, A. de; Koning, J.J. de; Daanen, H.A.M.

    2011-01-01

    Reliable continuous core temperature measurement is of major importance for monitoring patients. The zero heat flux method (ZHF) can potentially fulfil the requirements of non-invasiveness, reliability and short delay time that current measurement methods lack. The purpose of this study was to

  15. Wall temperature measurements using a thermal imaging camera with temperature-dependent emissivity corrections

    International Nuclear Information System (INIS)

    McDaid, Chloe; Zhang, Yang

    2011-01-01

    A methodology is presented whereby the relationship between temperature and emissivity for fused quartz has been used to correct the temperature values of a quartz impingement plate detected by an SC3000 thermal imaging camera. The methodology uses an iterative method using the initial temperature (obtained by assuming a constant emissivity) to find the emissivity values which are then put into the thermal imaging software and used to find the subsequent temperatures, which are used to find the emissivities, and so on until converged. This method is used for a quartz impingement plate that has been heated under various flame conditions, and the results are compared. Radiation losses from the plate are also calculated, and it is shown that even a slight change in temperature greatly affects the radiation loss. It is a general methodology that can be used for any wall material whose emissivity is a function of temperature

  16. Soil temperature variability in complex terrain measured using fiber-optic distributed temperature sensing

    Science.gov (United States)

    Soil temperature (Ts) exerts critical controls on hydrologic and biogeochemical processes but magnitude and nature of Ts variability in a landscape setting are rarely documented. Fiber optic distributed temperature sensing systems (FO-DTS) potentially measure Ts at high density over a large extent. ...

  17. Methods and Systems for Measurement and Estimation of Normalized Contrast in Infrared Thermography

    Science.gov (United States)

    Koshti, Ajay M. (Inventor)

    2017-01-01

    Methods and systems for converting an image contrast evolution of an object to a temperature contrast evolution and vice versa are disclosed, including methods for assessing an emissivity of the object; calculating an afterglow heat flux evolution; calculating a measurement region of interest temperature change; calculating a reference region of interest temperature change; calculating a reflection temperature change; calculating the image contrast evolution or the temperature contrast evolution; and converting the image contrast evolution to the temperature contrast evolution or vice versa, respectively.

  18. Comparison of two surface temperature measurement using thermocouples and infrared camera

    Directory of Open Access Journals (Sweden)

    Michalski Dariusz

    2017-01-01

    Full Text Available This paper compares two methods applied to measure surface temperatures at an experimental setup designed to analyse flow boiling heat transfer. The temperature measurements were performed in two parallel rectangular minichannels, both 1.7 mm deep, 16 mm wide and 180 mm long. The heating element for the fluid flowing in each minichannel was a thin foil made of Haynes-230. The two measurement methods employed to determine the surface temperature of the foil were: the contact method, which involved mounting thermocouples at several points in one minichannel, and the contactless method to study the other minichannel, where the results were provided with an infrared camera. Calculations were necessary to compare the temperature results. Two sets of measurement data obtained for different values of the heat flux were analysed using the basic statistical methods, the method error and the method accuracy. The experimental error and the method accuracy were taken into account. The comparative analysis showed that although the values and distributions of the surface temperatures obtained with the two methods were similar but both methods had certain limitations.

  19. X3 expansion tube driver gas spectroscopy and temperature measurements

    Science.gov (United States)

    Parekh, V.; Gildfind, D.; Lewis, S.; James, C.

    2017-11-01

    The University of Queensland's X3 facility is a large, free-piston driven expansion tube used for super-orbital and high Mach number scramjet aerothermodynamic studies. During recent development of new scramjet test flow conditions, experimentally measured shock speeds were found to be significantly lower than that predicted by initial driver performance calculations. These calculations were based on ideal, isentropic compression of the driver gas and indicated that loss mechanisms, not accounted for in the preliminary analysis, were significant. The critical determinant of shock speed is peak driver gas sound speed, which for a given gas composition depends on the peak driver gas temperature. This temperature may be inaccurately estimated if an incorrect fill temperature is assumed, or if heat losses during driver gas compression are significant but not accounted for. For this study, the ideal predicted peak temperature was 3750 K, without accounting for losses. However, a much lower driver temperature of 2400 K is suggested based on measured experimental shock speeds. This study aimed to measure initial and peak driver gas temperatures for a representative X3 operating condition. Examination of the transient temperatures of the driver gas and compression tube steel wall during the initial fill process showed that once the filling process was complete, the steady-state driver gas temperature closely matched the tube wall temperature. Therefore, while assuming the gas is initially at the ambient laboratory temperature is not a significant source of error, it can be entirely mitigated by simply monitoring tube wall temperature. Optical emission spectroscopy was used to determine the driver gas spectra after diaphragm rupture; the driver gas emission spectrum exhibited a significant continuum radiation component, with prominent spectral lines attributed to contamination of the gas. A graybody approximation of the continuum suggested a peak driver gas temperature of

  20. Determination of absorption coefficients of glasses at high tempera-tures, by measuring the thermal emission

    NARCIS (Netherlands)

    Loenen, E.; Van der Tempel, L.

    1996-01-01

    An experimental setup built in 1995 measures the spectral absorptioncoefficient of glass as a function of temperature and wavelength bythe emissive method. The setup was improved, as well as the softwarefor processing the measurement data. The measurement results of quartzwere validated by

  1. Brightness-temperature retrival methods in synthetic aperture radiometers

    OpenAIRE

    Corbella Sanahuja, Ignasi; Torres Torres, Francisco; Camps Carmona, Adriano José; Duffo Ubeda, Núria; Vall-Llossera Ferran, Mercedes Magdalena

    2009-01-01

    Bightness-temperature retrieval techniques for synthetic aperture radiometers are reviewed. Three different approaches to combine measured visibility and antenna temperatures, along with instrument characterization data, into a general equation to invert are presented. Discretization and windowing techniques are briefly discussed, and formulas for reciprocal grids using rectangular and hexagonal samplings are given. Two known techniques are used to invert the equation, namel...

  2. Self-calibrated active pyrometer for furnace temperature measurements

    Science.gov (United States)

    Woskov, Paul P.; Cohn, Daniel R.; Titus, Charles H.; Surma, Jeffrey E.

    1998-01-01

    Pyrometer with a probe beam superimposed on its field-of-view for furnace temperature measurements. The pyrometer includes a heterodyne millimeter/sub-millimeter-wave or microwave receiver including a millimeter/sub-millimeter-wave or microwave source for probing. The receiver is adapted to receive radiation from a surface whose temperature is to be measured. The radiation includes a surface emission portion and a surface reflection portion which includes the probe beam energy reflected from the surface. The surface emission portion is related to the surface temperature and the surface reflection portion is related to the emissivity of the surface. The simultaneous measurement of surface emissivity serves as a real time calibration of the temperature measurement. In an alternative embodiment, a translatable base plate and a visible laser beam allow slow mapping out of interference patterns and obtaining peak values therefor. The invention also includes a waveguide having a replaceable end portion, an insulating refractory sleeve and/or a source of inert gas flow. The pyrometer may be used in conjunction with a waveguide to form a system for temperature measurements in a furnace. The system may employ a chopper or alternatively, be constructed without a chopper. The system may also include an auxiliary reflector for surface emissivity measurements.

  3. Standard Test Method for Normal Spectral Emittance at Elevated Temperatures

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    1972-01-01

    1.1 This test method describes a highly accurate technique for measuring the normal spectral emittance of electrically conducting materials or materials with electrically conducting substrates, in the temperature range from 600 to 1400 K, and at wavelengths from 1 to 35 μm. 1.2 The test method requires expensive equipment and rather elaborate precautions, but produces data that are accurate to within a few percent. It is suitable for research laboratories where the highest precision and accuracy are desired, but is not recommended for routine production or acceptance testing. However, because of its high accuracy this test method can be used as a referee method to be applied to production and acceptance testing in cases of dispute. 1.3 The values stated in SI units are to be regarded as the standard. The values in parentheses are for information only. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this stan...

  4. Subassembly faults diagnostic of an LMFBR type reactor by the measurement of temperature noise

    International Nuclear Information System (INIS)

    Kokorev, B.V.; Palkin, I.I.; Turchin, N.M.; Pallagi, D.; Horanyi, S.

    1979-09-01

    The subassembly faults detection possibility by temperature noise analysis of an LMFBR is described. The paper contains the results of diagnostical examinations obtained on electrically heated NaK test rigs. On the basis of these results the measurement of temperature noise RMS value seems to be a practicable method to detect local blockages in an early phase. (author)

  5. Mold temperature measurement for glass-pressing processes

    International Nuclear Information System (INIS)

    Holman, R.A.

    1985-01-01

    The largest use of radiation thermometers within Corning Glass Works is for mold temperature measurement for the glass-pressing process. Pressing television panels at today's high quality would be very difficult without a mold temperature measurement system and the computer manipulation of the quality control data to supervise the mold temperature control loop. The most critical part of a television panel is the inside surface curvature. The ideal surface is usually defined as a spherical surface. The tolerance for a normal TV panel is +-0.30 mm (+-0.012 in.). High resolution display panels are more critical, having a dimensional tolerance only one half as large as TV panels. Panel curvature is a direct (but negative) function of mold temperature. Every 1 0 C increase in mold temperature results in the panel center being 0.025 mm (0.001 in.) shorter (flatter). Random dimensional variations within a panel take up most of the dimensional tolerance. The result is that each mold is controlled to its own individual temperature set point, +-1 0 C. Hot panel and cold panel curvature measurements are correlated by a process computer and used to update the mold temperature set points. The same computer adjusts the mold cooling air to maintain the required mold temperatures. From the temperature measurement standpoint, the significant problem is the changing emissivity of the mold surface when the mold is new or reconditioned. The selection of a radiation thermometer with a short wavelength was an obvious choice to minimize the effect of emissivity variations

  6. [Multispectral Radiation Algorithm Based on Emissivity Model Constraints for True Temperature Measurement].

    Science.gov (United States)

    Liang, Mei; Sun, Xiao-gang; Luan, Mei-sheng

    2015-10-01

    Temperature measurement is one of the important factors for ensuring product quality, reducing production cost and ensuring experiment safety in industrial manufacture and scientific experiment. Radiation thermometry is the main method for non-contact temperature measurement. The second measurement (SM) method is one of the common methods in the multispectral radiation thermometry. However, the SM method cannot be applied to on-line data processing. To solve the problems, a rapid inversion method for multispectral radiation true temperature measurement is proposed and constraint conditions of emissivity model are introduced based on the multispectral brightness temperature model. For non-blackbody, it can be drawn that emissivity is an increasing function in the interval if the brightness temperature is an increasing function or a constant function in a range and emissivity satisfies an inequality of emissivity and wavelength in that interval if the brightness temperature is a decreasing function in a range, according to the relationship of brightness temperatures at different wavelengths. The construction of emissivity assumption values is reduced from multiclass to one class and avoiding the unnecessary emissivity construction with emissivity model constraint conditions on the basis of brightness temperature information. Simulation experiments and comparisons for two different temperature points are carried out based on five measured targets with five representative variation trends of real emissivity. decreasing monotonically, increasing monotonically, first decreasing with wavelength and then increasing, first increasing and then decreasing and fluctuating with wavelength randomly. The simulation results show that compared with the SM method, for the same target under the same initial temperature and emissivity search range, the processing speed of the proposed algorithm is increased by 19.16%-43.45% with the same precision and the same calculation results.

  7. In-core moderator temperature measurement within candu reactors

    Science.gov (United States)

    Sion, N.

    1983-03-01

    The temperature profile of the D 2O moderator inside a CANDU (Canada Deuterium Uranium) reactor, within the calandria vessel, was measured by means of a specially instrumented probe introduced within the core. Measurements were made under steady and transient reactor conditions using two different sensors, viz. resistance temperature detectors (RTD) and type K chromel-alumel thermocouples. The results established the feasibility of in-core moderatortemperature measurement and indicated that the thermocouples used were relatively not affected by the intense radiation fields thus producing more accurate data.

  8. Space potential, temperature, and density profile measurements on RENTOR

    International Nuclear Information System (INIS)

    Schoch, P.M.

    1983-05-01

    Radial profiles of the space potential, electron temperature, and density have been measured on RENTOR with a heavy-ion-beam probe. The potential profile has been compared to predictions from a stochastic magnetic field fluctuation theory, using the measured temperature and density profiles. The comparison shows strong qualitative agreement in that the potential is positive and the order of T/sub e//e. There is some quantitative disagreement in that the measured radial electric fields are somewhat smaller than the theoretical predictions. To facilitate this comparison, a detailed analysis of the possible errors has been completed

  9. Weathering: methods and techniques to measure

    Science.gov (United States)

    Lopez-Arce, P.; Zornoza-Indart, A.; Alvarez de Buergo, M.; Fort, R.

    2012-04-01

    Surface recession takes place when weathered material is removed from the rocks. In order to know how fast does weathering and erosion occur, a review of several methods, analyses and destructive and non-destructive techniques to measure weathering of rocks caused by physico-chemical changes that occur in bedrocks due to salt crystallization, freezing-thaw, thermal shock, influence of water, wind, temperature or any type of environmental agent leading to weathering processes and development of soils, in-situ in the field or through experimental works in the laboratory are addressed. From micro-scale to macro-scale, from the surface down to more in depth, several case studies on in-situ monitoring of quantification of decay on soils and rocks from natural landscapes (mountains, cliffs, caves, etc) or from urban environment (foundations or facades of buildings, retaining walls, etc) or laboratory experimental works, such as artificial accelerated ageing tests (a.a.e.e.) or durability tests -in which one or more than one weathering agents are selected to assess the material behaviour in time and in a cyclic way- performed on specimens of these materials are summarised. Discoloration, structural alteration, precipitation of weathering products (mass transfer), and surface recession (mass loss) are all products of weathering processes. Destructive (SEM-EDX, optical microscopy, mercury intrusion porosimetry, drilling resistance measurement, flexural and compression strength) and Non-destructive (spectrophotocolorimetry, 3D optical surface roughness, Schmidt hammer rebound tester, ultrasound velocity propagation, Nuclear Magnetic Resonance NMR, X ray computed micro-tomography or CT-scan, geo-radar differential global positioning systems) techniques and characterization analyses (e.g. water absorption, permeability, open porosity or porosity accessible to water) to assess their morphological, physico-chemical, mechanical and hydric weathering; consolidation products or

  10. Innovative Instrumentation and Analysis of the Temperature Measurement for High Temperature Gasification

    Energy Technology Data Exchange (ETDEWEB)

    Seong W. Lee

    2005-10-01

    The objectives of this project during this semi-annual reporting period are to test the effects of coating layer of the thermal couple on the temperature measurement and to screen out the significant factors affecting the temperature reading under different operational conditions. The systematic tests of the gasifier simulator on the high velocity oxygen fuel (HVOF) spray coated thermal couples were completed in this reporting period. The comparison tests of coated and uncoated thermal couples were conducted under various operational conditions. The temperature changes were recorded and the temperature differences were calculated to describe the thermal spray coating effect on the thermal couples. To record the temperature data accurately, the computerized data acquisition system (DAS) was adopted to the temperature reading. The DAS could record the data with the accuracy of 0.1 C and the recording parameters are configurable. In these experiments, DAS was set as reading one data for every one (1) minute. The operational conditions are the combination of three parameters: air flow rate, water/ammonia flow rate and the amount of fine dust particles. The results from the temperature readings show the temperature of uncoated thermal couple is uniformly higher than that of coated thermal couple for each operational condition. Analysis of Variances (ANOVA) was computed based on the results from systematic tests to screen out the significant factors and/or interactions. The temperature difference was used as dependent variable and three operational parameters (i.e. air flow rate, water/ammonia flow rate and amount of fine dust particle) were used as independent factors. The ANOVA results show that the operational parameters are not the statistically significant factors affecting the temperature readings which indicate that the coated thermal couple could be applied to temperature measurement in gasifier. The actual temperature reading with the coated thermal couple in

  11. Measurement of critical temperature as a function of field

    Science.gov (United States)

    McInturff, A. D.; Ishibashi, K.; Heard, G. D.

    The critical temperature has been measured for various magnet conductors as a function of the perpendicular applied magnetic field. The isothermal environment was provided by a variable temperature cryostat which fits into the bore of a 10 telsa solenoid. The temperature gradient across the sample volume was measured to be less than 25 millikelvins. The superconducting to normal state transition was measured resistively, using sample current densities from 0.01 to 2 A cm -2. The maximum applied magnetic field was 10 T and varied less than 0.5% in the sample volume. The critical transport current range of the samples measured from tens to thousands of amperes in the presence of a 10 T perpendicular magnetic field at 4.2 K.

  12. Measurement of rock properties at elevated pressures and temperatures

    International Nuclear Information System (INIS)

    Pincus, H.J.; Hoskins, E.R.

    1985-01-01

    The papers in this volume were presented at an ASTM symposium held on 20 June 1983 in conjunction with the 24th Annual Rock Mechanics Symposium at Texas A and M University, College Station, TX. The purpose of these papers is to present recent developments in the measurement of rock properties at elevated pressures and temperatures, and to examine and interpret the data produced by such measurement. The need for measuring rock properties at elevated pressures and temperatures has become increasingly important in recent years. Location and design of nuclear waste repositories, development of geothermal energy sites, and design and construction of deep excavations for civil, military, and mining engineering require significantly improved capabilities for measuring rock properties under conditions substantially different from those prevailing in most laboratory and in situ work. The development of high-pressure, high-temperature capabilities is also significant for the analysis of tectonic processes

  13. A development of ultrasonic measurement at high temperature and the investigation of the ability of ultrasonic testing at room temperature

    International Nuclear Information System (INIS)

    Arakawa, T.; Yoshikawa, K.

    1988-01-01

    In order to evaluate the reliability of structures, it is important to develop the quantitative evaluation of natural cracks by ultrasonic examination. With the high accurate estimation of crack size and fracture toughness, which may be changeable in the case used at high temperature for long time, the simulation of crack extension can be possible by using the fracture mechanism. In the consideration of the predicted measuring errors, however, it will meet expectations to monitor the crack extension in operation. The high temperature probe was developed for this purpose and also for the use to estimate the degradation of materials. This high temperature probe demonstrated endurance at high temperature for long time without any coolant. This report summarizes the development of this high temperature probe in combination with the discussion of detectability using hydrogen-induced cracks by ultrasonic testing, from which the suitable test method can be selected not to miss the harmful cracks estimated from the fracture

  14. Handbook of nuclear safeguards measurement methods

    International Nuclear Information System (INIS)

    Rogers, D.R.

    1983-09-01

    This handbook is intended to be a guide to the selection of methods for meeting specific measurement requirements. The information was compiled from a survey of production facilities, the literature, and current exchange programs. The survey included bulk measurements, chemical assay, sampling techniques, isotopic measurements, passive NDA, and active NDA

  15. Methods of measurement in epidemiology: sedentary behaviour

    OpenAIRE

    Atkin, Andrew James; Gorely, Trish; Clemes, Stacy A; Yates, Thomas; Edwardson, Charlotte; Brage, Soren; Salmon, Jo; Marshall, Simon J; Biddle, Stuart JH

    2012-01-01

    Background: Research examining sedentary behaviour as a potentially independent risk factor for chronic disease morbidity and mortality has expanded rapidly in recent years. Methods: We present a narrative overview of the sedentary behaviour measurement literature. Subjective and objective methods of measuring sedentary behaviour suitable for use in population-based research with children and adults are examined. The validity and reliability of each method is considered, gaps in...

  16. Temperature profiles of different cooling methods in porcine pancreas procurement.

    Science.gov (United States)

    Weegman, Bradley P; Suszynski, Thomas M; Scott, William E; Ferrer Fábrega, Joana; Avgoustiniatos, Efstathios S; Anazawa, Takayuki; O'Brien, Timothy D; Rizzari, Michael D; Karatzas, Theodore; Jie, Tun; Sutherland, David E R; Hering, Bernhard J; Papas, Klearchos K

    2014-01-01

    Porcine islet xenotransplantation is a promising alternative to human islet allotransplantation. Porcine pancreas cooling needs to be optimized to reduce the warm ischemia time (WIT) following donation after cardiac death, which is associated with poorer islet isolation outcomes. This study examines the effect of four different cooling Methods on core porcine pancreas temperature (n = 24) and histopathology (n = 16). All Methods involved surface cooling with crushed ice and chilled irrigation. Method A, which is the standard for porcine pancreas procurement, used only surface cooling. Method B involved an intravascular flush with cold solution through the pancreas arterial system. Method C involved an intraductal infusion with cold solution through the major pancreatic duct, and Method D combined all three cooling Methods. Surface cooling alone (Method A) gradually decreased core pancreas temperature to procurement, but incorporating an intraductal infusion (Method C) rapidly reduced core temperature 15-20 °C within the first 2 min of cooling. Combining all methods (Method D) was the most effective at rapidly reducing temperature and providing sustained cooling throughout the duration of procurement, although the recorded WIT was not different between Methods (P = 0.36). Histological scores were different between the cooling Methods (P = 0.02) and the worst with Method A. There were differences in histological scores between Methods A and C (P = 0.02) and Methods A and D (P = 0.02), but not between Methods C and D (P = 0.95), which may highlight the importance of early cooling using an intraductal infusion. In conclusion, surface cooling alone cannot rapidly cool large (porcine or human) pancreata. Additional cooling with an intravascular flush and intraductal infusion results in improved core porcine pancreas temperature profiles during procurement and histopathology scores. These data may also have implications on human pancreas procurement as use of an

  17. A Temperature Sensor using a Silicon-on-Insulator (SOI) Timer for Very Wide Temperature Measurement

    Science.gov (United States)

    Patterson, Richard L.; Hammoud, Ahmad; Elbuluk, Malik; Culley, Dennis E.

    2008-01-01

    A temperature sensor based on a commercial-off-the-shelf (COTS) Silicon-on-Insulator (SOI) Timer was designed for extreme temperature applications. The sensor can operate under a wide temperature range from hot jet engine compartments to cryogenic space exploration missions. For example, in Jet Engine Distributed Control Architecture, the sensor must be able to operate at temperatures exceeding 150 C. For space missions, extremely low cryogenic temperatures need to be measured. The output of the sensor, which consisted of a stream of digitized pulses whose period was proportional to the sensed temperature, can be interfaced with a controller or a computer. The data acquisition system would then give a direct readout of the temperature through the use of a look-up table, a built-in algorithm, or a mathematical model. Because of the wide range of temperature measurement and because the sensor is made of carefully selected COTS parts, this work is directly applicable to the NASA Fundamental Aeronautics/Subsonic Fixed Wing Program--Jet Engine Distributed Engine Control Task and to the NASA Electronic Parts and Packaging (NEPP) Program. In the past, a temperature sensor was designed and built using an SOI operational amplifier, and a report was issued. This work used an SOI 555 timer as its core and is completely new work.

  18. Relationship between transepidermal water loss and temperature of the measuring probe.

    Science.gov (United States)

    Thoma, S; Welzel, J; Wilhelm, K P

    1997-02-01

    Transepidermal water loss (TEWL) is an important parameter for the determination of skin barrier function. The open chamber method has been established as the technique of choice in most dermatological laboratories for measurements of TEWL. However, the influence of the probe temperature on TEWL measurements has been the subject of recent controversial debates. In this study the relationship between TEWL measured with the Tewameter and temperature of the measuring probe was therefore investigated by comparing two different measuring techniques. For one measurement, the probe was kept at room temperature (20°C) and for the other one, the probe was preheated to the actual temperature of the measuring object before obtaining the values. Measurements were performed on evaporative standards (EvSs) and healthy individuals. For the EvSs, semipermeable membranes were pulled over a petri dish filled with water, which could be heated. TEWL values were found to depend on the temperature of the probe. TEWL values were higher when measured with the preheated probe. However, long-term measurements revealed that TEWL values measured with the unheated probe reached those higher TEWL values after approximately 8 min measuring time. The final TEWL value was reached after shorter intervals for the preheated probe compared to the unheated probe (2.5 min vs. 4 min) for some measurements. However, preheating of the probe resulted in greater variability of the measurement values. Therefore, measurements with a preheated Tewameter probe is not be recommended.

  19. Measuring method for optical fibre sensors

    NARCIS (Netherlands)

    Lammerink, Theodorus S.J.; Fluitman, J.H.J.

    1984-01-01

    A new measuring method for the signal amplitude in intensity modulating fibre optic sensors is described. A reference signal is generated in the time domain. The method is insensitive for the sensitivity fluctuations of the light transmitter and the light receiver. The method is experimentally

  20. Temperature Control of Autothermal Reformer System with Coefficient Diagram Method

    Science.gov (United States)

    Srisiriwat, N.; Wutthithanyawat, C.

    2017-10-01

    The objective of this paper is to design the autothermal reformer (ATR) temperature control by using a coefficient diagram method (CDM). The adiabatic temperature is a main controlled variable of the ATR which is a combination of endothermic and exothermic reactions. The simulation results of control parameters were calculated to maintain the ATR reaction temperature by manipulating air feed flow rate. In this work, two strategies of ATR temperature controller system with and without the feed temperature control of a preheater unit are compared to investigate the appropriate controller system when the change of surrounding temperature is considered as a key disturbance. The results showed that by using the CDM, the stability and robustness for controlling the ATR temperature system were considered to offer the proper control parameters and the designed temperature control of ATR system gave a good performance to maintain the controlled variables and reject the disturbance. Moreover, the ATR control system design with the feed temperature controller can compensate the surrounding temperature better than that without the feed temperature control.

  1. Fiber Bragg grating sensor for simultaneous measurement of temperature and force using polymer open loop

    Science.gov (United States)

    Huang, Yonglin; Zhang, Shiyan

    2014-07-01

    A fiber Bragg grating (FBG) sensor for simultaneous measurement of temperature and force is proposed and demonstrated. Where a part of uniform FBG (about one half length of an FBG) is attached on the polymer open loop, the FBG is divided into two parts which has an equal length. So the two parts can be regarded as two FBGs. Because of the difference of the Young's modulus and the thermal expansion coefficients for two parts of the FBG, the two Bragg reflection wavelengths are shift when the temperature and force are applied on the sensor. Simultaneous measurement of temperature and force is demonstrated experimentally. The experimental results show that the linear response to temperature and force are achieved. The value of applied temperature and force can be obtained from the two Bragg wavelength shift via the coefficient matrix. This study provides a simple and economical method to measure temperature and force simultaneously.

  2. Measurements of plasma temperature and electron density in laser ...

    Indian Academy of Sciences (India)

    using the Boltzmann plot method and the electron density is determined using the Saha–. Boltzmann equation method ... Laser-induced plasma; spectroscopy; plasma temperature; electron density. PACS Nos 52.50.Jm; 52.70. ... trace sample in any phase (solid, liquid and gas) with no or minimal sample prepa- ration [2–4].

  3. Determination of the yield locus by means of temperature measurement

    NARCIS (Netherlands)

    Banabic, D.; Huetink, Han

    2006-01-01

    The paper presents a theoretical background of the thermo-graphical method of determining the yield locus. The analytical expression of the temperature variation of the specimen deformed in the elastic state is determined starting from the first law of thermodynamics. The experimental method for

  4. Temperature, its measurement and control in industry - ITM '90

    International Nuclear Information System (INIS)

    Fischer, H.; Blieck, L.; Jescheck, M.; Neubert, W.; Kunze, D.

    1990-01-01

    The publication refers to the new VDE/VDI guideline 3511 and explains its practical intentions and implications by thoroughly discussing the applications of temperature sensors and their limits of use. The current state of the art in temperature measuring is fully described by the discussion of the new temperature scale introduced recently, the ITS '90. The authors of the book look in detail at the particular requirements and conditions of infrared measuring techniques using radiation thermometers as defined in DIN 5496, the applications of microprocessors (DIN-measuring-field-bus, etc.), time program emitters, and measuring transducers (EX ib, etc.). A full chapter has been devoted to the subject of surface temperature measurement. Examples referring to practical applications in industry serve as an introduction to thermal control engineering, in particular with infrared sensors, for processes such as thermal forming. New, optical thermometers for the low temperature range have been given much attention. An annex presents comprehensive tables for calculation and conversion of thermal quantities. (orig./HP) With 192 figs., 134 refs [de

  5. High-temperature absorbed dose measurements in the megagray range

    International Nuclear Information System (INIS)

    Balian, P.; Ardonceau, J.; Zuppiroli, L.

    1988-01-01

    Organic conductors of the tetraselenotetracene family have been tested as ''high-temperature'' absorbed dose dosimeters. They were heated up to 120 0 C and irradiated at this temperature with 1-MeV electrons in order to simulate, in a short time, a much longer γ-ray irradiation. The electric resistance increase of the crystal can be considered a good measurement of the absorbed dose in the range 10 6 Gy to a few 10 8 Gy and presumably one order of magnitude more. This dosimeter also permits on-line (in-situ) measurements of the absorbed dose without removing the sensor from the irradiation site. The respective advantages of organic and inorganic dosimeters at these temperature and dose ranges are also discussed. In this connection, we outline new, but negative, results concerning the possible use of silica as a high-temperature, high-dose dosimeter. (author)

  6. Gas Temperature Measurement in a Glow Discharge Plasma

    Science.gov (United States)

    Sloneker, Kenneth; Podder, Nirmol; McCurdy, William E.; Shi, Shi

    2009-10-01

    In this study a relatively inexpensive quartz protected thermocouple is used to measure the gas temperature in the positive column of a glow discharge plasma. For simplicity a K-type thermocouple is used to interpret the gas temperature from the sensor voltage at pressures from 0.5 Torr to 15 Torr and discharge currents from 5 mA to 120 mA. Gas temperature is investigated as a function of the gas pressure at fixed discharge currents and as a function of discharge current at fixed gas pressures in three different gas species (Ar, N2, and He). An infinite cylinder model is used to compute the average gas temperature of the discharge from joule heating and gas thermal conductivity. The model and measurement data agree within 1% to 10% depending on plasma parameters. Data for all three gases have a similar quasi-linear increasing error as compared to the model.

  7. Measuring Skin Temperatures with the IASI Hyperspectral Mission

    Science.gov (United States)

    Safieddine, S.; George, M.; Clarisse, L.; Clerbaux, C.

    2017-12-01

    Although the role of satellites in observing the variability of the Earth system has increased in recent decades, remote-sensing observations are still underexploited to accurately assess climate change fingerprints, in particular temperature variations. The IASI - Flux and Temperature (IASI-FT) project aims at providing new benchmarks for temperature observations using the calibrated radiances measured twice a day at any location by the IASI thermal infrared instrument on the suite of MetOp satellites (2006-2025). The main challenge is to achieve the accuracy and stability needed for climate studies, particularly that required for climate trends. Time series for land and sea skin surface temperatures are derived and compared with in situ measurements and atmospheric reanalysis. The observed trends are analyzed at seasonal and regional scales in order to disentangle natural (weather/dynamical) variability and human-induced climate forcings.

  8. Computation and measurement of a plasma temperature using the emission lines of copper

    International Nuclear Information System (INIS)

    Sassi, M.; Pierre, L.; Benard, J.; Cahen, C.

    1994-01-01

    The use of copper emission lines for temperature measurement in plasma environments is presented. The article features the problems encountered with such a measurement method when used in a stationary and a non-stationary plasma. In the first case, we were concerned with the temperature measurements in the jet of a 2 MW industrial plasma torche. The plasma was stationary and the measurement geometry allowed the use of the Abel inversion method to recover the temperature profiles in the plasma jet. The limitations of the measurement method on the jet boundaries as well as the cooling of the plasma by entrainement of cold air are discussed. In the second case, a non-stationary plasma in a 1.4 MW electric furnace was studied. This study allowed to feature the role of the detector dynamics as related to the dynamics of the observed medium. The obtained measurements show only a probable temperature in the immediate proximity of the arc. Finally, we conclude on the validity of copper as a tracer and the use of spontaneous emission spectroscopy, and easy method to implement, compared to other sophisticated temperature measurement methods. (orig.)

  9. Measuring temperature using MRI: a powerful and versatile technique.

    Science.gov (United States)

    Turner, Robert; Streicher, Markus

    2012-02-01

    The Larmor frequency of water protons has reliably linear temperature dependence. Since this frequency shift is easily measurable using relatively simple MRI techniques, a remarkable opportunity arises for uniquely non-invasive and accurate temperature evaluation, deep within any water-containing object. Major applications are appearing in the field of image-guided surgery. The cutting-edge papers collected in this Special Issue demonstrate both the versatility and the power of MRI thermometry.

  10. Study of high-temperature superconductors by IBA methods

    Science.gov (United States)

    Chamera, D.; Suchańska, M.; Kozel, V. V.

    1992-02-01

    The ion-induced electron emission (IIEE) technique to study the high- Tc superconductor of the Y 1Ba 2Cu 3O7 - x type was used. The temperature dependence of the IIEE intensity I( T) within the range of 77-300 K was measured. The change of the shape of the I( T) curve at the point 140 K was observed. The ion-induced photon emission technique (IIPE) to study the same high-temperature superconductor was also used. The temperature dependence of the intensity of some emission lines (BaI — 455.5 nm, CuI — 324.7 nm, YI — 467.4 nm) was measured. No anomaly at the point 140 K was observed. The nonmonotonic dependence of the IIEE intensity on the target temperature and the behaviour of the temperature dependence of the IIPE intensity are discussed.

  11. Study of high-temperature superconductors by IBA methods

    Energy Technology Data Exchange (ETDEWEB)

    Chamera, D.; Suchanska, M. (Inst. of Physics, Pedagogical Univ., Kielce (Poland)); Kozel, V.V. (Dept. of Physics, State Univ., Donetsk (USSR))

    1992-02-01

    The ion-induced electron emission (IIEE) technique to study the high-Tc superconductor of the Y{sub 1}Ba{sub 2}Cu{sub 3}O{sub 7-x} type was used. The temperature dependence of the IIEE intensity I(T) within the range of 77-300 K was measured. The change of the shape of the I(T) curve at the point 140 K was observed. The ion-induced photon emission technique (IIPE) to study the same high-temperature superconductor was also used. The temperature dependence of the intensity of some emission lines (BaI - 455.5 nm, CuI - 324.7 nm, YI - 467.4 nm) was measured. No anomaly at the point 140 K was observed. The nonmonotonic dependence of the IIEE intensity on the target temperature and the behaviour of the temperature dependence of the IIPE intensity are discussed. (orig.).

  12. Enhanced Temperature Control Method Using ANFIS with FPGA

    Directory of Open Access Journals (Sweden)

    Chiung-Wei Huang

    2014-01-01

    Full Text Available Temperature control in etching process is important for semiconductor manufacturing technology. However, pressure variations in vacuum chamber results in a change in temperature, worsening the accuracy of the temperature of the wafer and the speed and quality of the etching process. This work develops an adaptive network-based fuzzy inference system (ANFIS using a field-programmable gate array (FPGA to improve the effectiveness. The proposed method adjusts every membership function to keep the temperature in the chamber stable. The improvement of the proposed algorithm is confirmed using a medium vacuum (MV inductively-coupled plasma- (ICP- type etcher.

  13. Thermal history sensors for non-destructive temperature measurements in harsh environments

    Energy Technology Data Exchange (ETDEWEB)

    Pilgrim, C. C. [Mechanical Engineering, Imperial College London, London, SW7 2AZ, UK and Sensor Coating Systems, Imperial Incubator, Bessemer Building, Level 1 and 2, Imperial College London, London SW7 2AZ (United Kingdom); Heyes, A. L. [Energy Technology and Innovation Initiative, University of Leeds, Leeds, LS2 9JT (United Kingdom); Feist, J. P. [Sensor Coating Systems, Imperial Incubator, Bessemer Building, Level 1 and 2, Imperial College London, London SW7 2AZ (United Kingdom)

    2014-02-18

    The operating temperature is a critical physical parameter in many engineering applications, however, can be very challenging to measure in certain environments, particularly when access is limited or on rotating components. A new quantitative non-destructive temperature measurement technique has been proposed which relies on thermally induced permanent changes in ceramic phosphors. This technique has several distinct advantages over current methods for many different applications. The robust ceramic material stores the temperature information allowing long term thermal exposures in harsh environment to be measured at a convenient time. Additionally, rare earth dopants make the ceramic phosphorescent so that the temperature information can be interpreted by automated interrogation of the phosphorescent light. This technique has been demonstrated by application of YAG doped with dysprosium and europium as coatings through the air-plasma spray process. Either material can be used to measure temperature over a wide range, namely between 300°C and 900°C. Furthermore, results show that the material records the peak exposure temperature and prolonged exposure at lower temperatures would have no effect on the temperature measurement. This indicates that these materials could be used to measure peak operating temperatures in long-term testing.

  14. Diamond micro-Raman thermometers for accurate gate temperature measurements

    Energy Technology Data Exchange (ETDEWEB)

    Simon, Roland B.; Pomeroy, James W.; Kuball, Martin [Center for Device Thermography and Reliability, H. H. Wills Physics Laboratory, University of Bristol, Tyndall Avenue, Bristol BS8 1TL (United Kingdom)

    2014-05-26

    Determining the peak channel temperature in AlGaN/GaN high electron mobility transistors and other devices with high accuracy is an important and challenging issue. A surface-sensitive thermometric technique is demonstrated, utilizing Raman thermography and diamond microparticles to measure the gate temperature. This technique enhances peak channel temperature estimation, especially when it is applied in combination with standard micro-Raman thermography. Its application to other metal-covered areas of devices, such as field plates is demonstrated. Furthermore, this technique can be readily applied to other material/device systems.

  15. Diamond micro-Raman thermometers for accurate gate temperature measurements

    International Nuclear Information System (INIS)

    Simon, Roland B.; Pomeroy, James W.; Kuball, Martin

    2014-01-01

    Determining the peak channel temperature in AlGaN/GaN high electron mobility transistors and other devices with high accuracy is an important and challenging issue. A surface-sensitive thermometric technique is demonstrated, utilizing Raman thermography and diamond microparticles to measure the gate temperature. This technique enhances peak channel temperature estimation, especially when it is applied in combination with standard micro-Raman thermography. Its application to other metal-covered areas of devices, such as field plates is demonstrated. Furthermore, this technique can be readily applied to other material/device systems.

  16. Study of Optical Fiber Sensors for Cryogenic Temperature Measurements

    Directory of Open Access Journals (Sweden)

    Veronica De Miguel-Soto

    2017-11-01

    Full Text Available In this work, the performance of five different fiber optic sensors at cryogenic temperatures has been analyzed. A photonic crystal fiber Fabry-Pérot interferometer, two Sagnac interferometers, a commercial fiber Bragg grating (FBG, and a π-phase shifted fiber Bragg grating interrogated in a random distributed feedback fiber laser have been studied. Their sensitivities and resolutions as sensors for cryogenic temperatures have been compared regarding their advantages and disadvantages. Additionally, the results have been compared with the given by a commercial optical backscatter reflectometer that allowed for distributed temperature measurements of a single mode fiber.

  17. Study of Optical Fiber Sensors for Cryogenic Temperature Measurements.

    Science.gov (United States)

    De Miguel-Soto, Veronica; Leandro, Daniel; Lopez-Aldaba, Aitor; Beato-López, Juan Jesus; Pérez-Landazábal, José Ignacio; Auguste, Jean-Louis; Jamier, Raphael; Roy, Philippe; Lopez-Amo, Manuel

    2017-11-30

    In this work, the performance of five different fiber optic sensors at cryogenic temperatures has been analyzed. A photonic crystal fiber Fabry-Pérot interferometer, two Sagnac interferometers, a commercial fiber Bragg grating (FBG), and a π-phase shifted fiber Bragg grating interrogated in a random distributed feedback fiber laser have been studied. Their sensitivities and resolutions as sensors for cryogenic temperatures have been compared regarding their advantages and disadvantages. Additionally, the results have been compared with the given by a commercial optical backscatter reflectometer that allowed for distributed temperature measurements of a single mode fiber.

  18. Temperature measurements in a wall stabilized steady flame using CARS

    KAUST Repository

    Sesha Giri, Krishna

    2017-01-05

    Flame quenching by heat loss to a surface continues to be an active area of combustion research. Close wall temperature measurements in an isothermal wall-stabilized flame are reported in this work. Conventional N-vibrational Coherent Anti-Stokes Raman Scattering (CARS) thermometry as close as 275 μm to a convex wall cooled with water has been carried out. The standard deviation of mean temperatures is observed to be ~6.5% for high temperatures (>2000K) and ~14% in the lower range (<500K). Methane/air and ethylene/air stoichiometric flames for various global strain rates based on exit bulk velocities are plotted and compared. CH* chemiluminescence is employed to determine the flame location relative to the wall. Flame locations are shown to move closer to the wall with increasing strain rates in addition to higher near-wall temperatures. Peak temperatures for ethylene are considerably higher (~250-300K) than peak temperatures for methane. Preheat zone profiles are similar for different strain rates across fuels. This work demonstrates close wall precise temperature measurments using CARS.

  19. Downhole temperature tool accurately measures well bore profile

    International Nuclear Information System (INIS)

    Cloud, W.B.

    1992-01-01

    This paper reports that an inexpensive temperature tool provides accurate temperatures measurements during drilling operations for better design of cement jobs, workovers, well stimulation, and well bore hydraulics. Valid temperature data during specific wellbore operations can improve initial job design, fluid testing, and slurry placement, ultimately enhancing well bore performance. This improvement applies to cement slurries, breaker activation for slurries, breaker activation for stimulation and profile control, and fluid rheological properties for all downhole operations. The temperature tool has been run standalone mounted inside drill pipe, on slick wire line and braided cable, and as a free-falltool. It has also been run piggyback on both directional surveys (slick line and free-fall) and standard logging runs. This temperature measuring system has been used extensively in field well bores to depths of 20,000 ft. The temperature tool is completely reusable in the field, ever similar to the standard directional survey tools used on may drilling rigs. The system includes a small, rugged, programmable temperature sensor, a standard body housing, various adapters for specific applications, and a personal computer (PC) interface

  20. Methods of measuring residual stresses in components

    International Nuclear Information System (INIS)

    Rossini, N.S.; Dassisti, M.; Benyounis, K.Y.; Olabi, A.G.

    2012-01-01

    Highlights: ► Defining the different methods of measuring residual stresses in manufactured components. ► Comprehensive study on the hole drilling, neutron diffraction and other techniques. ► Evaluating advantage and disadvantage of each method. ► Advising the reader with the appropriate method to use. -- Abstract: Residual stresses occur in many manufactured structures and components. Large number of investigations have been carried out to study this phenomenon and its effect on the mechanical characteristics of these components. Over the years, different methods have been developed to measure residual stress for different types of components in order to obtain reliable assessment. The various specific methods have evolved over several decades and their practical applications have greatly benefited from the development of complementary technologies, notably in material cutting, full-field deformation measurement techniques, numerical methods and computing power. These complementary technologies have stimulated advances not only in measurement accuracy and reliability, but also in range of application; much greater detail in residual stresses measurement is now available. This paper aims to classify the different residual stresses measurement methods and to provide an overview of some of the recent advances in this area to help researchers on selecting their techniques among destructive, semi destructive and non-destructive techniques depends on their application and the availabilities of those techniques. For each method scope, physical limitation, advantages and disadvantages are summarized. In the end this paper indicates some promising directions for future developments.

  1. Adaptive measurement method for miniature spectrometers used in cold environments.

    Science.gov (United States)

    Wang, Hangzhou; Nan, Liwen; Huang, Hui; Yang, Ping; Song, Hong; Han, Jiwan; Wu, Yuanqian; Yan, Tingting; Yuan, Zhuoli; Chen, Ying

    2017-10-01

    Adaptive measurement is a major concern when using miniature spectrometers in extreme environments, especially when the ambient temperatures and incident light intensities vary greatly. In this study, parameters, including the signal output and the relevant noise and signal-to-noise ratio (SNR) of a fiber optic spectrometry system composed of a photodiode array miniature spectrometer and external driver electronics were examined at multiple integration times from -50°C to 30°C, well below the specified operating temperature of this spectrometer. The relationships between those parameters and incident light level were also examined, at a single temperature of 0°C. Based on these examinations, temperature-induced biases in the linear operating range of the spectrometer were identified. Signal output and the relevant noise and SNR in response to different integration times, temperatures, and incident light levels were assessed separately. These assessments were then used to develop an adaptive measurement method for estimating the incident light level and setting up an optimal integration time for this spectrometer, while autonomously adapting the variation in the ambient temperature and incident light level simultaneously. This approach provides a general framework for developing an adaptive measurement algorithm for miniature spectrometers, which face tremendous variations in ambient temperature and incident light level.

  2. Measurement of QCD gluon propagator at finite temperature

    Energy Technology Data Exchange (ETDEWEB)

    Saito, Takuya [Hiroshima Univ., Information Media Center, Higashi-Hiroshima, Hiroshima (Japan); Nakamura, Atsushi [Yamagata Univ., Faculty of Education, Yamagata (Japan); Sakai, Sunao [Hiroshima Univ., Faculty of Integrated Arts and Sciences, Higashi-Hiroshima, Hiroshima (Japan)

    2002-09-01

    We calculate gluon screening masses in quenched SU(3) lattice gauge theory at finite temperature by using stochastic gauge fixing method. Although gluon propagator is gauge dependent quantity and is not physical object, we extract the screening masses of SU(3) gluon by the behaviour of gauge fixed gluon propagator. We examine temperature dependences of the electric and magnetic screening masses at the regions, T/T{sub c} = 1 - 6. We also check their gauge dependence is very small. (author)

  3. Contemporary methods of body composition measurement.

    Science.gov (United States)

    Fosbøl, Marie Ø; Zerahn, Bo

    2015-03-01

    Reliable and valid body composition assessment is important in both clinical and research settings. A multitude of methods and techniques for body composition measurement exist, all with inherent problems, whether in measurement methodology or in the assumptions upon which they are based. This review is focused on currently applied methods for in vivo measurement of body composition, including densitometry, bioimpedance analysis, dual-energy X-ray absorptiometry, computed tomography (CT), magnetic resonance techniques and anthropometry. Multicompartment models including quantification of trace elements by in vivo neutron activation analysis, which are regarded as gold standard methods, are also summarized. The choice of a specific method or combination of methods for a particular study depends on various considerations including accuracy, precision, subject acceptability, convenience, cost and radiation exposure. The relative advantages and disadvantages of each method are discussed with these considerations in mind. © 2014 Scandinavian Society of Clinical Physiology and Nuclear Medicine. Published by John Wiley & Sons Ltd.

  4. Contemporary methods of body composition measurement

    DEFF Research Database (Denmark)

    Fosbøl, Marie Ø; Zerahn, Bo

    2015-01-01

    Reliable and valid body composition assessment is important in both clinical and research settings. A multitude of methods and techniques for body composition measurement exist, all with inherent problems, whether in measurement methodology or in the assumptions upon which they are based. This re...... and radiation exposure. The relative advantages and disadvantages of each method are discussed with these considerations in mind........ This review is focused on currently applied methods for in vivo measurement of body composition, including densitometry, bioimpedance analysis, dual-energy X-ray absorptiometry, computed tomography (CT), magnetic resonance techniques and anthropometry. Multicompartment models including quantification of trace...

  5. Artefacts in intracavitary temperature measurements during regional hyperthermia

    International Nuclear Information System (INIS)

    Kok, H P; Berg, C A T Van den; Haaren, P M A Van; Crezee, J

    2007-01-01

    For adequate hyperthermia treatments, reliable temperature information during treatment is essential. During regional hyperthermia, temperature information is preferably obtained non-invasively from intracavitary or intraluminal measurements to avoid implant risks for the patient. However, for intracavitary or intraluminal thermometry optimal tissue contact is less natural as for invasive thermometry. In this study, the reliability of intraluminal/intracavitary measurements was examined in phantom experiments and in a numerical model for various extents of thermal contact between thermometry and the surroundings. Both thermocouple probes and fibre optic probes were investigated. Temperature rises after a 30 s power pulse of the 70 MHz AMC-4 hyperthermia system were measured in a tissue-equivalent phantom using a multisensor thermocouple probe placed centrally in a hollow tube. The tube was filled with (1) air (2) distilled water or (3) saline solution that mimics the properties of tissue, simulating situations with (1) bad thermal contact and no power dissipation in the tube (2) good thermal contact but no power dissipation or (3) good thermal contact and tissue representative power dissipation. For numerical simulations, a cylindrical symmetric model of a thermocouple probe or a fibre optic probe in a cavity was developed. The cavity was modelled as air, distilled water or saline solution. A generalised E-Field distribution was assumed, resulting in a power deposition. With this power deposition, the temperature rise after a 30 s power pulse was calculated. When thermal contact was bad (1), both phantom measurements and simulations with a thermocouple probe showed very high temperature rises (>0.5 0 C), which are artefacts due to self-heating of the thermocouple probe, since no power is dissipated in air. Simulations with a fibre optic probe showed almost no temperature rise when the cavity was filled with air. When thermal contact was good, but no power was

  6. Sensitive Dependence of Gibbs Measures at Low Temperatures

    Science.gov (United States)

    Coronel, Daniel; Rivera-Letelier, Juan

    2015-09-01

    The Gibbs measures of an interaction can behave chaotically as the temperature drops to zero. We observe that for some classical lattice systems there are interactions exhibiting a related phenomenon of sensitive dependence of Gibbs measures: An arbitrarily small perturbation of the interaction can produce significant changes in the low-temperature behavior of its Gibbs measures. For some one-dimensional XY models we exhibit sensitive dependence of Gibbs measures for a (nearest-neighbor) interaction given by a smooth function, and for perturbations that are small in the smooth category. We also exhibit sensitive dependence of Gibbs measures for an interaction on a classical lattice system with finite-state space. This interaction decreases exponentially as a function of the distance between sites; it is given by a Lipschitz continuous potential in the configuration space. The perturbations are small in the Lipschitz topology. As a by-product we solve some problems stated by Chazottes and Hochman.

  7. Simple method for highlighting the temperature distribution into a liquid sample heated by microwave power field

    International Nuclear Information System (INIS)

    Surducan, V.; Surducan, E.; Dadarlat, D.

    2013-01-01

    Microwave induced heating is widely used in medical treatments, scientific and industrial applications. The temperature field inside a microwave heated sample is often inhomogenous, therefore multiple temperature sensors are required for an accurate result. Nowadays, non-contact (Infra Red thermography or microwave radiometry) or direct contact temperature measurement methods (expensive and sophisticated fiber optic temperature sensors transparent to microwave radiation) are mainly used. IR thermography gives only the surface temperature and can not be used for measuring temperature distributions in cross sections of a sample. In this paper we present a very simple experimental method for temperature distribution highlighting inside a cross section of a liquid sample, heated by a microwave radiation through a coaxial applicator. The method proposed is able to offer qualitative information about the heating distribution, using a temperature sensitive liquid crystal sheet. Inhomogeneities as smaller as 1°-2°C produced by the symmetry irregularities of the microwave applicator can be easily detected by visual inspection or by computer assisted color to temperature conversion. Therefore, the microwave applicator is tuned and verified with described method until the temperature inhomogeneities are solved

  8. Methods of Measurement in epidemiology: Sedentary Behaviour

    Science.gov (United States)

    Atkin, Andrew J; Gorely, Trish; Clemes, Stacy A; Yates, Thomas; Edwardson, Charlotte; Brage, Soren; Salmon, Jo; Marshall, Simon J; Biddle, Stuart JH

    2012-01-01

    Background Research examining sedentary behaviour as a potentially independent risk factor for chronic disease morbidity and mortality has expanded rapidly in recent years. Methods We present a narrative overview of the sedentary behaviour measurement literature. Subjective and objective methods of measuring sedentary behaviour suitable for use in population-based research with children and adults are examined. The validity and reliability of each method is considered, gaps in the literature specific to each method identified and potential future directions discussed. Results To date, subjective approaches to sedentary behaviour measurement, e.g. questionnaires, have focused predominantly on TV viewing or other screen-based behaviours. Typically, such measures demonstrate moderate reliability but slight to moderate validity. Accelerometry is increasingly being used for sedentary behaviour assessments; this approach overcomes some of the limitations of subjective methods, but detection of specific postures and postural changes by this method is somewhat limited. Instruments developed specifically for the assessment of body posture have demonstrated good reliability and validity in the limited research conducted to date. Miniaturization of monitoring devices, interoperability between measurement and communication technologies and advanced analytical approaches are potential avenues for future developments in this field. Conclusions High-quality measurement is essential in all elements of sedentary behaviour epidemiology, from determining associations with health outcomes to the development and evaluation of behaviour change interventions. Sedentary behaviour measurement remains relatively under-developed, although new instruments, both objective and subjective, show considerable promise and warrant further testing. PMID:23045206

  9. Estimation of piping temperature fluctuations based on external strain measurements

    International Nuclear Information System (INIS)

    Morilhat, P.; Maye, J.P.

    1993-01-01

    Due to the difficulty to carry out measurements at the inner sides of nuclear reactor piping subjected to thermal transients, temperature and stress variations in the pipe walls are estimated by means of external thermocouples and strain-gauges. This inverse problem is solved by spectral analysis. Since the wall harmonic transfer function (response to a harmonic load) is known, the inner side signal will be obtained by convolution of the inverse transfer function of the system and of the strain measurement enables detection of internal temperature fluctuations in a frequency range beyond the scope of the thermocouples. (authors). 5 figs., 3 refs

  10. Temperature dependence of the Brewer global UV measurements

    Science.gov (United States)

    Fountoulakis, Ilias; Redondas, Alberto; Lakkala, Kaisa; Berjon, Alberto; Bais, Alkiviadis F.; Doppler, Lionel; Feister, Uwe; Heikkila, Anu; Karppinen, Tomi; Karhu, Juha M.; Koskela, Tapani; Garane, Katerina; Fragkos, Konstantinos; Savastiouk, Volodya

    2017-11-01

    Spectral measurements of global UV irradiance recorded by Brewer spectrophotometers can be significantly affected by instrument-specific optical and mechanical features. Thus, proper corrections are needed in order to reduce the associated uncertainties to within acceptable levels. The present study aims to contribute to the reduction of uncertainties originating from changes in the Brewer internal temperature, which affect the performance of the optical and electronic parts, and subsequently the response of the instrument. Until now, measurements of the irradiance from various types of lamps at different temperatures have been used to characterize the instruments' temperature dependence. The use of 50 W lamps was found to induce errors in the characterization due to changes in the transmissivity of the Teflon diffuser as it warms up by the heat of the lamp. In contrast, the use of 200 or 1000 W lamps is considered more appropriate because they are positioned at longer distances from the diffuser so that warming is negligible. Temperature gradients inside the instrument can cause mechanical stresses which can affect the instrument's optical characteristics. Therefore, during the temperature-dependence characterization procedure warming or cooling must be slow enough to minimize these effects. In this study, results of the temperature characterization of eight different Brewer spectrophotometers operating in Greece, Finland, Germany and Spain are presented. It was found that the instruments' response changes differently in different temperature regions due to different responses of the diffusers' transmittance. The temperature correction factors derived for the Brewer spectrophotometers operating at Thessaloniki, Greece, and Sodankylä, Finland, were evaluated and were found to remove the temperature dependence of the instruments' sensitivity.

  11. Load forecasting method considering temperature effect for distribution network

    Directory of Open Access Journals (Sweden)

    Meng Xiao Fang

    2016-01-01

    Full Text Available To improve the accuracy of load forecasting, the temperature factor was introduced into the load forecasting in this paper. This paper analyzed the characteristics of power load variation, and researched the rule of the load with the temperature change. Based on the linear regression analysis, the mathematical model of load forecasting was presented with considering the temperature effect, and the steps of load forecasting were given. Used MATLAB, the temperature regression coefficient was calculated. Using the load forecasting model, the full-day load forecasting and time-sharing load forecasting were carried out. By comparing and analyzing the forecast error, the results showed that the error of time-sharing load forecasting method was small in this paper. The forecasting method is an effective method to improve the accuracy of load forecasting.

  12. Experimental study of the use of probe methods for diagnostic of low-temperature magnetized plasma

    OpenAIRE

    Zanáška, Michal

    2015-01-01

    The ball-pen probe is a relatively new diagnostic method, that has been designed for direct measurement of plasma potential in magnetized plasmas. Nowadays, it is routinely used at several high-temperature plasma devices in Europe and it has been tested also in conditions of low-temperature plasma, which are substantially different from that of high-temperature plasma. The measurements performed so far showed, that ball-pen probe could be used also in low-temperature plasma. However, more mea...

  13. Determination of magnetic characteristics of nanoparticles by low-temperature calorimetry methods

    Energy Technology Data Exchange (ETDEWEB)

    Ugulava, A.; Toklikishvili, Z. [Department of Physics, I.Javakhishvili Tbilisi State University,I.Chavchavadze av. 3, 0179 Tbilisi, Georgia (United States); Chkhaidze, S., E-mail: simon.chkhaidze@tsu.ge [Department of Physics, I.Javakhishvili Tbilisi State University,I.Chavchavadze av. 3, 0179 Tbilisi, Georgia (United States); Kekutia, Sh. [V. Chavchanidze Institute of Cybernetics, at the Technical State University, S. Euli str. 5, 0186 Tbilisi, Georgia (United States)

    2017-05-15

    At low temperatures, the heat capacity of a superparamagnetic “ideal gas” determined by magnetic degrees of freedom can greatly exceed the lattice heat capacity. It is shown that in the presence of an external magnetic field, the temperature dependence of the magnetic part of the heat capacity has two maxima. The relations between the temperature at which these maxima are achieved, the magnetic moment of the nanoparticles and the magnetic anisotropy constant have been obtained. Measuring the heat capacity maxima temperatures by low-temperature calorimetry methods and using the obtained relations, we can obtain the numerical values both of the magnetic moment of nanoparticles and the magnetic anisotropy constants.

  14. The Role of ADC-Based Thermometry in Measuring Brain Intraventricular Temperature in Children.

    Science.gov (United States)

    Wagner, Matthias W; Stern, Steven E; Oshmyansky, Alexander; Huisman, Thierry A G M; Poretti, Andrea

    2016-05-01

    To determine the feasibility of apparent diffusion coefficient (ADC)-based thermometry to assess intraventricular temperature in children. ADC maps were generated from diffusion tensor imaging data, which were acquired with diffusion gradients along 20 noncollinear directions using a b-value of 1000 s/mm(2) . The intraventricular temperature was calculated based on intraventricular ADC values and the mode method as previously reported. The calculated intraventricular temperature was validated with an estimated brain temperature based on temporal artery temperature measurements. We included 120 children in this study (49 females, 71 males, mean age 6.63 years), 15 consecutive children for each of the following age groups: 0-1, 1-2, 2-4, 4-6, 6-8, 8-10, 10-14, and 14-18 years. Forty-three children had a normal brain MRI and 77 children had an abnormal brain scan. Polynomial fitting to the temperature distribution and subsequent calculation of mode values was performed. A correlation coefficient and a coefficient of determination were calculated between ADC calculated temperatures and estimated brain temperatures. Linear regression analysis was performed to investigate the two temperature measures. ADC-based intraventricular temperatures ranged between 31.5 and 39.6 °C, although estimated brain temperatures ranged between 36.3 and 38.1 °C. The difference between the temperatures is larger for children with more than 8,000 voxels within the lateral ventricles compared to children with less than 8,000 voxels. The correlation coefficient between ADC-based temperatures and the estimated brain temperatures is .1, the respective R(2) is .01 indicating that 1% of the changes in estimated brain temperatures are attributable to corresponding changes in ADC-based temperature measurements (P = .275). ADC-based thermometry has limited application in the pediatric population mainly due to a small ventricular size. Copyright © 2015 by the American Society of Neuroimaging.

  15. INNOVATIVE INSTRUMENTATION AND ANALYSIS OF THE TEMPERATURE MEASUREMENT FOR HIGH TEMPERATURE GASIFICATION

    Energy Technology Data Exchange (ETDEWEB)

    Seong W. Lee

    2005-04-01

    The systematic tests of the gasifier simulator on the ultrasonic vibration application for cleaning method were completed in this reporting period. Within the systematic tests on the ultrasonic vibration application, the ambient temperature and high temperature status condition were tested separately. The sticky dirt on the thermocouple tip was simulated by the cement-covered layer on the thermocouple tip. At the ambient temperature status, four (4) factors were considered as the input factors affecting the response variable of peeling off rate. The input factors include the shape of the cement-covered layer (thickness and length), the ultrasonic vibration output power, and application time. At the high temperature tests, four (4) different environments were considered as the experimental parameters including air flow supply, water and air supply environment, water/air/fine dust particle supply, and air/water/ammonia/fine dust particle supply environment. The factorial design method was used in the experiment design with twelve (12) data sets of readings. Analysis of Variances (ANOVA) was applied to the results from systematic tests. The ANOVA results show that the thickness and length of the cement-covered layer have the significant impact on the peeling off rate of ultrasonic vibration application at the ambient temperature environment. For the high temperature tests, the different environments do not seem to have significant impact on the temperature changes. These results may indicate that the ultrasonic vibration is one of best cleaning methods for the thermocouple tip.

  16. Novel Zero-Heat-Flux Deep Body Temperature Measurement in Lower Extremity Vascular and Cardiac Surgery.

    Science.gov (United States)

    Mäkinen, Marja-Tellervo; Pesonen, Anne; Jousela, Irma; Päivärinta, Janne; Poikajärvi, Satu; Albäck, Anders; Salminen, Ulla-Stina; Pesonen, Eero

    2016-08-01

    The aim of this study was to compare deep body temperature obtained using a novel noninvasive continuous zero-heat-flux temperature measurement system with core temperatures obtained using conventional methods. A prospective, observational study. Operating room of a university hospital. The study comprised 15 patients undergoing vascular surgery of the lower extremities and 15 patients undergoing cardiac surgery with cardiopulmonary bypass. Zero-heat-flux thermometry on the forehead and standard core temperature measurements. Body temperature was measured using a new thermometry system (SpotOn; 3M, St. Paul, MN) on the forehead and with conventional methods in the esophagus during vascular surgery (n = 15), and in the nasopharynx and pulmonary artery during cardiac surgery (n = 15). The agreement between SpotOn and the conventional methods was assessed using the Bland-Altman random-effects approach for repeated measures. The mean difference between SpotOn and the esophageal temperature during vascular surgery was+0.08°C (95% limit of agreement -0.25 to+0.40°C). During cardiac surgery, during off CPB, the mean difference between SpotOn and the pulmonary arterial temperature was -0.05°C (95% limits of agreement -0.56 to+0.47°C). Throughout cardiac surgery (on and off CPB), the mean difference between SpotOn and the nasopharyngeal temperature was -0.12°C (95% limits of agreement -0.94 to+0.71°C). Poor agreement between the SpotOn and nasopharyngeal temperatures was detected in hypothermia below approximately 32°C. According to this preliminary study, the deep body temperature measured using the zero-heat-flux system was in good agreement with standard core temperatures during lower extremity vascular and cardiac surgery. However, agreement was questionable during hypothermia below 32°C. Copyright © 2016 Elsevier Inc. All rights reserved.

  17. Image-Method Gain Measurement With Mismatch

    Science.gov (United States)

    Lee, Richard Q.; Baddour, Maurice F.

    1988-01-01

    New formula accounts for multiple reflections. In image method antenna placed facing its image in reflecting plane. Power transmitted by antenna and portion of transmitted power received by antenna after reflection from plane measured at various distances R.

  18. Real-time measurement of aerosol particle concentration at high temperatures; Hiukkaspitoisuuden reaaliaikainen mittaaminen korkeassa laempoetilassa

    Energy Technology Data Exchange (ETDEWEB)

    Keskinen, J.; Hautanen, J.; Laitinen, A. [Tampere Univ. of Technology (Finland). Physics

    1997-10-01

    The aim of this project is to develop a new method for continuous aerosol particle concentration measurement at elevated temperatures (up to 800-1000 deg C). The measured property of the aerosol particles is the so called Fuchs surface area. This quantity is relevant for diffusion limited mass transfer to particles. The principle of the method is as follows. First, aerosol particles are charged electrically by diffusion charging process. The charging takes place at high temperature. After the charging, aerosol is diluted and cooled. Finally, aerosol particles are collected and the total charge carried by the aerosol particles is measured. Particle collection and charge measurement take place at low temperature. Benefits of this measurement method are: particles are charged in-situ, charge of the particles is not affected by the temperature and pressure changes after sampling, particle collection and charge measurement are carried out outside the process conditions, and the measured quantity is well defined. The results of this study can be used when the formation of the fly ash particles is studied. Another field of applications is the study and the development of gasification processes. Possibly, the method can also be used for the monitoring the operation of the high temperature particle collection devices. (orig.)

  19. Measurements of the gap/displacement and development of the ultrasonic temperature measuring system applied to severe accidents research

    International Nuclear Information System (INIS)

    Koo, Kil Mo; Kang, Kyung Ho; Cho, Young Ro; Park, Rae Jun; Kim, Sang Baik; Sim, Chul Moo

    2001-02-01

    This report, in order to measure quantitative LAVA experimental results, focuses on measuring the gap formed on the lower head vessel using a ultrasonic pulse echo method and neutron radiography, measuring displacement of the lower head vessel using capacitance method, building a measuring system and developing high temperature measurement system using ultrasonic method. The scope of gap measurement and system development using the ultrasonic method is 2-dimensional image processing using tomographical B scan method and 2- and 3-dimensional image processing using C scan methods based on the one dimensional time domain A scan signal. For some test specimen, the gap size is quantitative represented apply C scan methods. The important ultrasonic image processing technique is on the development of accurate position control system. The requirements of the position control system are a contact technique on the test specimen and a fine moving technique. Since the specimen is hemispherical, the contact technique is very difficult. Therefore, the gap measurement using the ultrasonic pulse echo method was applied developing the position controlling scanner system. Along with the ultrasonic method, neutron radiography method using KAERI's neutron source was attempted 4 times and the results are compared. The fine displacement of the hemispherical specimen was measured using a capacitive displacement sensor. The requirements for this measuring technique are fixing of the capacitance sensor to the experimental facilities and a remote control position varying system. This remote control position varying system was manufactured with a electrical motor. The development of a high temperature measuring system using a ultrasonic method the second year plan, is performed with developing a sensor which can measure up to 2300 deg C

  20. Liquid level measurement on coolant pipeline using Raman distributed temperature sensor

    International Nuclear Information System (INIS)

    Kasinathan, M.; Sosamma, S.; Babu Rao, C.; Murali, N.; Jayakumar, T.

    2011-01-01

    Optical fibre based Raman Distributed Temperature Sensor (RDTS) has been widely used for temperature monitoring in oil pipe line, power cable and environmental monitoring. Recently it has gained importance in nuclear reactor owing to its advantages like continuous, distributed temperature monitoring and immunity from electromagnetic interference. It is important to monitor temperature based level measurement in sodium capacities and in coolant pipelines for Fast Breeder Reactor (FBR). This particular application is used for filling and draining sodium in storage tank of sodium circuits of Fast breeder reactor. There are different conventional methods to find out the sodium level in the storage tank of sodium cooled reactors. They are continuous level measurement and discontinuous level measurement. For continuous level measurement, mutual inductance type level probes are used. The disadvantage of using this method is it needs a temperature compensation circuit. For discontinuous level measurement, resistance type discontinuous level probe and mutual inductance type discontinuous level probe are used. In resistance type discontinuous level probe, each level needs a separate probe. To overcome these disadvantages, RDTS is used for level measurement based distributed temperature from optical fibre as sensor. The feasibility of using RDTS for measurement of temperature based level measurement sensor is studied using a specially designed test set-up and using hot water, instead of sodium. The test set-up consist of vertically erected Stainless Steel (SS) pipe of length 2m and diameter 10cm, with provision for filling and draining out the liquid. Bare graded index multimode fibre is laid straight along the length of the of the SS pipe. The SS pipe is filled with hot water at various levels. The hot water in the SS pipe is maintained at constant temperature by insulating the SS pipe. The temperature profile of the hot water at various levels is measured using RDTS. The

  1. Core-temperature sensor ingestion timing and measurement variability.

    Science.gov (United States)

    Domitrovich, Joseph W; Cuddy, John S; Ruby, Brent C

    2010-01-01

    Telemetric core-temperature monitoring is becoming more widely used as a noninvasive means of monitoring core temperature during athletic events. To determine the effects of sensor ingestion timing on serial measures of core temperature during continuous exercise. Crossover study. Outdoor dirt track at an average ambient temperature of 4.4°C ± 4.1°C and relative humidity of 74.1% ± 11.0%. Seven healthy, active participants (3 men, 4 women; age  =  27.0 ± 7.5 years, height  =  172.9 ± 6.8 cm, body mass  =  67.5 ± 6.1 kg, percentage body fat  =  12.7% ± 6.9%, peak oxygen uptake [Vo(2peak)]  =  54.4 ± 6.9 mL•kg⁻¹•min⁻¹) completed the study. Participants completed a 45-minute exercise trial at approximately 70% Vo(2peak). They consumed core-temperature sensors at 24 hours (P1) and 40 minutes (P2) before exercise. Core temperature was recorded continuously (1-minute intervals) using a wireless data logger worn by the participants. All data were analyzed using a 2-way repeated-measures analysis of variance (trial × time), Pearson product moment correlation, and Bland-Altman plot. Fifteen comparisons were made between P1 and P2. The main effect of time indicated an increase in core temperature compared with the initial temperature. However, we did not find a main effect for trial or a trial × time interaction, indicating no differences in core temperature between the sensors (P1  =  38.3°C ± 0.2°C, P2  =  38.3°C ± 0.4°C). We found no differences in the temperature recordings between the 2 sensors. These results suggest that assumed sensor location (upper or lower gastrointestinal tract) does not appreciably alter the transmission of reliable and repeatable measures of core temperature during continuous running in the cold.

  2. Development of a Temperature Compensation Method for the Radiation Detector of a Radiation Monitoring System

    International Nuclear Information System (INIS)

    Park, Ki Sun; Jeong, Jae Jun

    2014-01-01

    Americium, which has been used to date for the assessment of integrity and efficiency of the radiation detector, was replaced by an LED to resolve these problems. However, while americium is negligibly affected by the temperature and the surrounding environment, the LED is significantly affected by the temperature. Accordingly, a method to resolve this issue is needed. In this study, a temperature-compensated preamplifier was developed, where the reference luminous output of the LED affecting on temperatures is constantly maintained to be uniform by controlling the current and the PMT gain difference is compensated by controlling the high voltage from a comparison of the input and output pulse values. After setting up the experimental apparatus, the validity of the preamplifier was verified through radiation measurement experiments where the temperature was varied from 10 .deg. C to 60 .deg. C. The output signals of the scintillation detector of the RMS are intricately related to the temperature. In particular, the output signal value is greatly affected by the thermal equilibrium in the process of converting electro-magnetic energy to light energy and back to electrical energy. To compensate for these temperature characteristics, the americium standard source is used, since it is not affected by temperature, or a temperature compensation method using an LED optical-pulse as a reference value is used. In this study, instead of the americium standard source, an LED and a temperature detector were used to develop a temperature compensation device for the PMT amplification factor. To verify the validity of the developed method, a reference temperature of 25 .deg. C and a reference current of a 60μA LED optical-pulse were used to experimentally measure the radiation while varying the temperature from 10 .deg. C to 60 .deg. C using a Cs-137 standard source. The experiment results showed that the measurement error is reduced by approximately 83.3 % by using the

  3. In situ infrared emission spectroscopy for quantitative gas-phase measurement under high temperature reaction conditions: an analytical method for methane by means of an innovative small-volume flowing cell.

    Science.gov (United States)

    Usseglio, Sandro; Thorshaug, Knut; Karlsson, Arne; Dahl, Ivar M; Nielsen, Claus J; Jens, Klaus-J; Tangstad, Elisabeth

    2010-02-01

    We have used infrared emission spectroscopy (IRES) in order to perform in situ studies under flowing gas-phase conditions. When the small-volume cell developed herein is used, we can (1) observe emission spectra from a hot gas-phase sample having an effective volume much less than one milliliter, (2) observe spectra of typical molecular species present, and (3) observe spectra of the more important molecular species down to below 10% and in some cases even as low as 1%. In addition, an analytical method has been derived in order to conduct quantitative studies under typical reaction conditions. We show that simplifications can be made in the data acquisition and handling for a direct linear correlation between band intensity and concentration with only simple background correction. The practical lower limit for methane in the present setup is approximately 0.5-1% v/v depending on the selected temperature. Our data were collected at 500, 600, and 700 degrees C, respectively. The major features of the present cell design are fairly simple and basically formed by a quartz tube (outer diameter=6 mm, inner diameter=4 mm) inside a metal pipe and two tubular ceramic heaters. This simple setup has advantages and attractive features that have extended the application of IRES to new fields and, in particular, for in situ studies of hydrocarbon reactions at different residence times at high temperature.

  4. Constituent Ion Temperatures Measured in the Topside Ionosphere

    Science.gov (United States)

    Hsu, C. T.; Heelis, R. A.

    2017-12-01

    Plasma temperatures in the ionosphere are associated with both the dynamics and structure of the neutral and charge particles. The temperatures are determined by solar energy inputs and energy exchange between charged particles and neutrals. Previous observations show that during daytime the O+ temperature is generally higher when the fractional contribution of H+ to the plasma is high. Further simulations confirm that the daytime heat balance between the H+ and O+ always keeps the H+ at a temperature higher than the O+. In addition the plasma transport parallel and perpendicular to the magnetic field influences the plasma temperature through adiabatic heating and cooling effects. These processes are also important during the nighttime, when the source of photoionization is absent. In this work we examine a more sophisticated analysis procedure to extract individual mass dependent ion temperature and apply it on the DMSP F15 RPA measurements. The result shows that the daytime TH+ is a few hundred degrees higher than the TO+ and the nighttime temperature difference between TH+ and TO+ is indicative of mass dependent adiabatic heating and cooling processes across the equatorial region.

  5. On Possibility of Detonation Products Temperature Measurements of Emulsion Explosives

    Directory of Open Access Journals (Sweden)

    Silvestrov V. V.

    2014-10-01

    Full Text Available The new view on the structure of the radiance signal recorded by optical pyrometer and the preliminary results of brightness detonation temperature of the emulsion explosive are presented. The structure of an optical signal observed is typical for the heterogeneous explosives. First, there is the short temperature spike to 2500 ÷ 3300 K connecting with a formation of “hot spots” assembly that fire the matrix capable of exothermal reaction. Then the relaxation of radiance to equilibrium level is observed that corresponds to brightness temperature 1840 ÷ 2260 K of explosion products at detonation pressure 1 ÷ 11 GPa. Experimental results are compared with the calculations of other authors. The detonation temperature of the investigated explosive is measured for the first time.

  6. CARS Temperature Measurements in a Combustion-Heated Supersonic Jet

    Science.gov (United States)

    Tedder, S. A.; Danehy, P. M.; Magnotti, G.; Cutler, A. D.

    2009-01-01

    Measurements were made in a combustion-heated supersonic axi-symmetric free jet from a nozzle with a diameter of 6.35 cm using dual-pump Coherent Anti-Stokes Raman Spectroscopy (CARS). The resulting mean and standard deviation temperature maps are presented. The temperature results show that the gas temperature on the centerline remains constant for approximately 5 nozzle diameters. As the heated gas mixes with the ambient air further downstream the mean temperature decreases. The standard deviation map shows evidence of the increase of turbulence in the shear layer as the jet proceeds downstream and mixes with the ambient air. The challenges of collecting data in a harsh environment are discussed along with influences to the data. The yield of the data collected is presented and possible improvements to the yield is presented are discussed.

  7. A lidar system for measuring atmospheric pressure and temperature profiles

    Science.gov (United States)

    Schwemmer, Geary K.; Dombrowski, Mark; Korb, C. Laurence; Milrod, Jeffry; Walden, Harvey

    1987-01-01

    The design and operation of a differential absorption lidar system capable of remotely measuring the vertical structure of tropospheric pressure and temperature are described. The measurements are based on the absorption by atmospheric oxygen of the spectrally narrowband output of two pulsed alexandrite lasers. Detailed laser output spectral characteristics, which are critical to successful lidar measurements, are presented. Spectral linewidths of 0.026 and 0.018 per cm for the lasers were measured with over 99.99 percent of the energy contained in three longitudinal modes.

  8. A new method to measure vaginal sensibility

    NARCIS (Netherlands)

    Lakeman, M. M. E.; Laan, E.; Vaart, C. H.; Roovers, J. P.

    2010-01-01

    Vaginal surgery may affect sexual function both positively and negatively. Possibly, negative consequences of surgical interventions on sexuality may be caused by reduced sensibility of the vaginal wall. To develop a new method to measure vaginal sensibility. We developed a technique to measure the

  9. Electrical measurement of absolute temperature and temperature transients in a buried nanostructure under ultrafast optical heating

    Science.gov (United States)

    Yang, H. F.; Hu, X. K.; Liebing, N.; Böhnert, T.; Costa, J. D.; Tarequzzaman, M.; Ferreira, R.; Sievers, S.; Bieler, M.; Schumacher, H. W.

    2017-06-01

    We report absolute temperature measurements in a buried nanostructure with a sub-nanosecond temporal resolution. For this purpose, we take advantage of the temperature dependence of the resistance of a magnetic tunnel junction (MTJ) as detected by a fast sampling oscilloscope. After calibrating the measurement setup using steady-state electric heating, we are able to quantify temperature changes in the MTJ induced by femtosecond optical heating of the metal contact lying several 100 nm above the MTJ. We find that a femtosecond pulse train with an average power of 400 mW and a repetition rate of 76 MHz leads to a constant temperature increase of 80 K and a temporally varying temperature change of 2 K in the MTJ. The maximum temperature change in the MTJ occurs 4 ns after the femtosecond laser pulses hit the metal contact, which is supported by simulations. Our work provides a scheme to quantitatively study local temperatures in nanoscale structures and might be important for the testing of nanoscale thermal transport simulations.

  10. Comparison of MODIS-derived land surface temperature with air temperature measurements

    Science.gov (United States)

    Georgiou, Andreas; Akçit, Nuhcan

    2017-09-01

    Air surface temperature is an important parameter for a wide range of applications such as agriculture, hydrology and climate change studies. Air temperature data is usually obtained from measurements made in meteorological stations, providing only limited information about spatial patterns over wide areas. The use of remote sensing data can help overcome this problem, particularly in areas with low station density, having the potential to improve the estimation of air surface temperature at both regional and global scales. Land Surface (skin) Temperatures (LST) derived from Moderate Resolution Imaging Spectroradiometer (MODIS) sensor aboard the Terra and Aqua satellite platforms provide spatial estimates of near-surface temperature values. In this study, LST values from MODIS are compared to groundbased near surface air (Tair) measurements obtained from 14 observational stations during 2011 to 2015, covering coastal, mountainous and urban areas over Cyprus. Combining Terra and Aqua LST-8 Day and Night acquisitions into a mean monthly value, provide a large number of LST observations and a better overall agreement with Tair. Comparison between mean monthly LSTs and mean monthly Tair for all sites and all seasons pooled together yields a very high correlation and biases. In addition, the presented high standard deviation can be explained by the influence of surface heterogeneity within MODIS 1km2 grid cells, the presence of undetected clouds and the inherent difference between LST and Tair. However, MODIS LST data proved to be a reliable proxy for surface temperature and mostly for studies requiring temperature reconstruction in areas with lack of observational stations.

  11. A study on the applicability of implantable microchip transponders for body temperature measurements in pigs

    DEFF Research Database (Denmark)

    Lohse, Louise; Uttenthal, Åse; Enøe, Claes

    2010-01-01

    Background The applicability of an electronic monitoring system using microchip transponders for measurement of body temperatures was tested in 6-week-old conventional Danish weaners infected with classical swine fever virus (CSFV). Subcutaneous tissue temperatures obtained by the implantable...... transponders were compared with rectal temperatures, recorded by a conventional digital thermometer. Methods In a preliminary study, transponders were inserted subcutaneously at 6 different positions of the body of 5 pigs. The transponders positioned by the ear base provided the best correlation to rectal......C lower than the rectal temperature. However, a simple linear relationship between the measures of the two methods was found. Conclusions Our study showed that the tested body monitoring system may represent a promising tool to obtain an approximate correlate of body temperatures in groups of pigs...

  12. A New Method to Measure Vaginal Sensibility

    NARCIS (Netherlands)

    Lakeman, M. M. E.; Laan, E.; Vaart, C. H.; Roovers, J. P.

    2010-01-01

    Introduction: Vaginal surgery may affect sexual function both positively and negatively. Possibly, negative consequences of surgical interventions on sexuality may be caused by reduced sensibility of the vaginal wall. Aims: To develop a new method to measure vaginal sensibility. Methods: We

  13. Power raise through improved reactor inlet header temperature measurement at Bruce A Nuclear Generation Station

    International Nuclear Information System (INIS)

    Basu, S.; Bruggemn, D.

    1997-01-01

    Reactor Inlet Header (RIH) temperature has become a factor limiting the performance of the Ontario Hydro Bruce A units. Specifically, the RIH temperature is one of several parameters that is preventing the Bruce A units from returning to 94% power operation. RIH temperature is one of several parameters which affect the critical heat flux in the reactor channel, and hence the integrity of the fuel. Ideally, RIH temperature should be lowered, but this cannot be done without improving the heat transfer performance of the boilers and feedwater pre-heaters. Unfortunately, the physical performance of the boilers and pre-heaters has decayed and continues to decay over time and as a result the RIH temperature has been rising and approaching its defined limit. With an understanding of the current RIH temperature measurement loop and methods available to improve it, a solution to reduce the measurement uncertainty is presented

  14. Infrared thermography for temperature measurement and non-destructive testing.

    Science.gov (United States)

    Usamentiaga, Rubén; Venegas, Pablo; Guerediaga, Jon; Vega, Laura; Molleda, Julio; Bulnes, Francisco G

    2014-07-10

    The intensity of the infrared radiation emitted by objects is mainly a function of their temperature. In infrared thermography, this feature is used for multiple purposes: as a health indicator in medical applications, as a sign of malfunction in mechanical and electrical maintenance or as an indicator of heat loss in buildings. This paper presents a review of infrared thermography especially focused on two applications: temperature measurement and non-destructive testing, two of the main fields where infrared thermography-based sensors are used. A general introduction to infrared thermography and the common procedures for temperature measurement and non-destructive testing are presented. Furthermore, developments in these fields and recent advances are reviewed.

  15. Estimating local heat transfer coefficients from thin wall temperature measurements

    Science.gov (United States)

    Gazizov, I. M.; Davletshin, I. A.; Paereliy, A. A.

    2017-09-01

    An approach to experimental estimation of local heat transfer coefficient on a plane wall has been described. The approach is based on measurements of heat-transfer fluid and wall temperatures during some certain time of wall cooling. The wall was a thin plate, a printed circuit board, made of composite epoxy material covered with a copper layer. The temperature field can be considered uniform across the plate thickness when heat transfer is moderate and thermal resistance of the plate in transversal direction is low. This significantly simplifies the heat balance written for the wall sections that is used to estimate the heat transfer coefficient. The copper layer on the plate etched to form a single strip acted as resistance thermometers that measured the local temperature of the wall.

  16. Measured energy savings from using night temperature setback

    International Nuclear Information System (INIS)

    Szydlowski, R.F.; Wrench, L.E.; O'Neill, P.J.

    1993-01-01

    The measured energy savings resulting from using night temperature setback in typical light-construction wooden office buildings was determined. Researchers installed monitoring equipment in a six-building sample of two-story wooden buildings at Fort Devens, Massachusetts. Data obtained during both single-setting and night-setback operating modes were used to develop models of each building's heat consumption as a function of the difference between inside and outside temperature. These models were used to estimate seasonal savings that could be obtained from the use of night-setback thermostat control. The measured savings in heating energy from using night temperature setback for the six Fort Devens buildings ranged from 14% to 25%; the mean savings was 19.2%. Based on an energy cost of $0.65/therm of natural gas, the estimated average cost savings of using automatic setback thermostats in these buildings is $780 per year per building

  17. Infrared Thermography for Temperature Measurement and Non-Destructive Testing

    Science.gov (United States)

    Usamentiaga, Rubèn; Venegas, Pablo; Guerediaga, Jon; Vega, Laura; Molleda, Julio; Bulnes, Francisco G.

    2014-01-01

    The intensity of the infrared radiation emitted by objects is mainly a function of their temperature. In infrared thermography, this feature is used for multiple purposes: as a health indicator in medical applications, as a sign of malfunction in mechanical and electrical maintenance or as an indicator of heat loss in buildings. This paper presents a review of infrared thermography especially focused on two applications: temperature measurement and non-destructive testing, two of the main fields where infrared thermography-based sensors are used. A general introduction to infrared thermography and the common procedures for temperature measurement and non-destructive testing are presented. Furthermore, developments in these fields and recent advances are reviewed. PMID:25014096

  18. Infrared Thermography for Temperature Measurement and Non-Destructive Testing

    Directory of Open Access Journals (Sweden)

    Rubén Usamentiaga

    2014-07-01

    Full Text Available The intensity of the infrared radiation emitted by objects is mainly a function of their temperature. In infrared thermography, this feature is used for multiple purposes: as a health indicator in medical applications, as a sign of malfunction in mechanical and electrical maintenance or as an indicator of heat loss in buildings. This paper presents a review of infrared thermography especially focused on two applications: temperature measurement and non-destructive testing, two of the main fields where infrared thermography-based sensors are used. A general introduction to infrared thermography and the common procedures for temperature measurement and non-destructive testing are presented. Furthermore, developments in these fields and recent advances are reviewed.

  19. Parametrization of optimum filter passbands for rotational Raman temperature measurements.

    Science.gov (United States)

    Hammann, Eva; Behrendt, Andreas

    2015-11-30

    We revisit the methodology of rotational Raman temperature measurements covering both lidar and non-range-resolved measurements, e.g., for aircraft control. The results of detailed optimization calculations are presented for the commonly used extraction of signals from the anti-Stokes branch. Different background conditions and realistic shapes of the filter transmission curves are taken into account. Practical uncertainties of the central passbands and widths are discussed. We found a simple parametrization for the optimum filter passband shifts depending on the atmospheric temperature range of interest and the background. The approximation errors of this parametrization are smaller than 2% for temperatures between 200 and 300 K and smaller than 4% between 180 and 200 K.

  20. Temperature measurement in cans of fuel rods and fuel rod simulators

    International Nuclear Information System (INIS)

    Tschoeke, H.; Moeller, R.

    1977-01-01

    On the sodium-cooled 19-rod cluster model for the SNR 300 the can wall temperature distributions of the non-uniformly cooled rods were measured with thermocouples mounted in outer grooves in the peripheral zone, permitting, in connection with Ni solder, a practically undisturbed measurement. For a more exact determination of the local surface temperature a calibration method, the so-called double-wall method, was developed and applied. The description of this calibration method and the experimental results achieved until now are presented. (orig./RW) [de

  1. A Measurement of the Cosmic Microwave Background Temperature ...

    Indian Academy of Sciences (India)

    tribpo

    •calibrating the contributions from the feed assembly and receiver. The thermodynamic temperature of the CMB is estimated to be. 3.45 ± 0.78 K. Key words. Cosmic microwave background — cosmology: observations. 1. Introduction. The spectrum of the cosmic microwave background (CMB) has been measured by the.

  2. Tokamak Plasmas: Measurement of temperature fluctuations and ...

    Indian Academy of Sciences (India)

    ... Refresher Courses · Symposia · Live Streaming. Home; Journals; Pramana – Journal of Physics; Volume 55; Issue 5-6. Tokamak Plasmas : Measurement of temperature fluctuations and anomalous transport in the SINP tokamak. R Kumar S K Saha. Contributed Papers Volume 55 Issue 5-6 November-December 2000 pp ...

  3. Measurements of plasma temperature and electron density in laser

    Indian Academy of Sciences (India)

    The temperature and electron density characterizing the plasma are measured by time-resolved spectroscopy of neutral atom and ion line emissions in the time window of 300–2000 ns. An echelle spectrograph coupled with a gated intensified charge coupled detector is used to record the plasma emissions.

  4. Measurements of plasma temperature and electron density in laser ...

    Indian Academy of Sciences (India)

    Home; Journals; Pramana – Journal of Physics; Volume 74; Issue 6. Measurements of plasma temperature and electron density in laser-induced copper plasma by time-resolved spectroscopy of neutral atom and ion emissions. V K Unnikrishnan Kamlesh Alti V B Kartha C Santhosh G P Gupta B M Suri. Research Articles ...

  5. Absolute method of measuring magnetic susceptibility

    Science.gov (United States)

    Thorpe, A.; Senftle, F.E.

    1959-01-01

    An absolute method of standardization and measurement of the magnetic susceptibility of small samples is presented which can be applied to most techniques based on the Faraday method. The fact that the susceptibility is a function of the area under the curve of sample displacement versus distance of the magnet from the sample, offers a simple method of measuring the susceptibility without recourse to a standard sample. Typical results on a few substances are compared with reported values, and an error of less than 2% can be achieved. ?? 1959 The American Institute of Physics.

  6. Method of orthogonally splitting imaging pose measurement

    Science.gov (United States)

    Zhao, Na; Sun, Changku; Wang, Peng; Yang, Qian; Liu, Xintong

    2018-01-01

    In order to meet the aviation's and machinery manufacturing's pose measurement need of high precision, fast speed and wide measurement range, and to resolve the contradiction between measurement range and resolution of vision sensor, this paper proposes an orthogonally splitting imaging pose measurement method. This paper designs and realizes an orthogonally splitting imaging vision sensor and establishes a pose measurement system. The vision sensor consists of one imaging lens, a beam splitter prism, cylindrical lenses and dual linear CCD. Dual linear CCD respectively acquire one dimensional image coordinate data of the target point, and two data can restore the two dimensional image coordinates of the target point. According to the characteristics of imaging system, this paper establishes the nonlinear distortion model to correct distortion. Based on cross ratio invariability, polynomial equation is established and solved by the least square fitting method. After completing distortion correction, this paper establishes the measurement mathematical model of vision sensor, and determines intrinsic parameters to calibrate. An array of feature points for calibration is built by placing a planar target in any different positions for a few times. An terative optimization method is presented to solve the parameters of model. The experimental results show that the field angle is 52 °, the focus distance is 27.40 mm, image resolution is 5185×5117 pixels, displacement measurement error is less than 0.1mm, and rotation angle measurement error is less than 0.15°. The method of orthogonally splitting imaging pose measurement can satisfy the pose measurement requirement of high precision, fast speed and wide measurement range.

  7. Implantable microchip transponders for body temperature measurements in pigs

    DEFF Research Database (Denmark)

    Lohse, Louise; Uttenthal, Åse; Enøe, Claes

    body temperature was tested, in order to evaluate the utility and reliability of this tool, in domestic pigs. The system is presently used and well optimized in small laboratory animals [1, 2]. We tested the microchip transponders during experimental infection of pigs with classical swine fever virus...... (CSFV), a viral infection, which can cause high fever in infected animals. Materials and Methods Implantable, programmable temperature transponders (IPTT-300TM) from Bio Medic Data System (Plexx, the Netherlands), designed for non-surgical implantation into animals, was tested in 30 weaner pigs. One...... methods showed that the peripheral subcutaneous body temperature recorded by a microchip transponder may be interesting as a monitoring tool in the clinical surveillance of the health status of domestic pigs. This technology has not at present the power to monitor individually sick pigs, where accurate...

  8. COMPANY PERFORMANCE MEASUREMENT AND REPORTING METHODS

    Directory of Open Access Journals (Sweden)

    Nicu Ioana Elena

    2012-12-01

    Full Text Available One of the priorities of economic research has been and remains the re-evaluation of the notion of performance and especially exploring and finding some indicators that would reflect as accurately as possible the subtleties of the economic entity. The main purpose of this paper is to highlight the main company performance measurement and reporting methods. Performance is a concept that raises many question marks concerning the most accurate or the best method of reporting the performance at the company level. The research methodology has aimed at studying the Romanian and foreign specialized literature dealing with the analyzed field, studying magazines specialized on company performance measurement. If the financial performance measurement indicators are considered to offer an accurate image of the situation of the company, the modern approach through non-financial indicators offers a new perspective upon performance measurement, which is based on simplicity. In conclusion, after the theoretical study, I have noticed that the methods of performance measurement, reporting and interpretation are various, the opinions regarding the best performance measurement methods are contradictive and the companies prefer resorting to financial indicators that still play a more important role in the consolidation of the company performance measurement than the non-financial indicators do.

  9. Measurements of Humidity and Temperature in the Marine Environment during the HEXOS Main Experiment

    NARCIS (Netherlands)

    Katsaros, K.B.; Cosmo, J. de; Lind, R.J.; Anderson, R.J.; Smith, S.D.; Kraan, R.; Oost, W.A.; Uhlig, K.; Mestayer, P.G.; Larsen, S.E.; Smith, M.H.; Leeuw, G. de

    1994-01-01

    Accurate measurement of fluctuations in temperature and humidity are needed for determination of the surface evaporation rate and the air-sea sensible heat flux using either the eddy correlation or inertial dissipation method for flux calculations. These measurements are difficult to make over the

  10. The use of spontaneous voltage fluctuations for the measurement of low temperatures

    NARCIS (Netherlands)

    Endt, P.M.

    1950-01-01

    A discussion is given of the method proposed by Lawson and Long to use the noise across a resistance as a measure of its absolute temperature. The statistical fluctuations in the measured noise are calculated, taking into account the bandwidth of the amplifier. This calculation shows that is not

  11. Optical Method for Detecting Displacements and Strains at Ultra-High Temperatures During Thermo-Mechanical Testing

    Science.gov (United States)

    Smith, Russell W. (Inventor); Rivers, H. Kevin (Inventor); Sikora, Joseph G. (Inventor); Roth, Mark C. (Inventor); Johnston, William M. (Inventor)

    2016-01-01

    An ultra-high temperature optical method incorporates speckle optics for sensing displacement and strain measurements well above conventional measurement techniques. High temperature pattern materials are used which can endure experimental high temperature environments while simultaneously having a minimum optical aberration. A purge medium is used to reduce or eliminate optical distortions and to reduce, and/or eliminate oxidation of the target specimen.

  12. Core Temperature Measurement During Submaximal Exercise: Esophageal, Rectal, and Intestinal Temperatures

    Science.gov (United States)

    Lee, Stuart M. C.; Williams, W. Jon; Schneider, Suzanne M.

    2000-01-01

    The purpose of this study was to determine if intestinal temperature (Tin) might be in acceptable alternative to esophageal (Tes) and rectal temperature (Trec) to assess thermoregulation during supine exercise. We hypothesized that Tin would have values similar to Tes and a response time similar to Trec, but the rate of temperature change across time would not be different between measurement sites. Seven subjects completed a continuous supine protocol of 20 min of rest, 20 min of cycle exercise at 40% peak oxygen consumption (VO2pk), 20 min of cycle exercise at 65% V02pk, and 20 min of recovery. Tes, Trec, and Tin were recorded each min throughout the test. Temperatures were not different after 20 min of rest, but Trec was less than the Tes and Tin at the end of the 40% and 65% VO2pk stages. After 20 min of recovery, Tes was less than either Trec or Tin, which were not different from each other. Time to threshold for increased temperature from rest was greater for Trec than Tes but not different from Tin. Time to reach peak temperature was greater for Tin and Trec than Tes. Similarly, time to a decrease in temperature after exercise was greater for Trec than Tes, but not different from Tin. The rate of temperature change from threshold to the end of the 40% VO2pk stage was not different between measurement sites. However, the rate of change during recovery was more negative for Tes than Tin and Trec, which were different from each other. Measurement of Tin may he an acceptable alternative to Tes and Trec with an understanding of its limitations.

  13. Carbon nanotubes and methods of forming same at low temperature

    Science.gov (United States)

    Biris, Alexandru S.; Dervishi, Enkeleda

    2017-05-02

    In one aspect of the invention, a method for growth of carbon nanotubes includes providing a graphitic composite, decorating the graphitic composite with metal nanostructures to form graphene-contained powders, and heating the graphene-contained powders at a target temperature to form the carbon nanotubes in an argon/hydrogen environment that is devoid of a hydrocarbon source. In one embodiment, the target temperature can be as low as about 150.degree. C. (.+-.5.degree. C.).

  14. New fitting method for temperature from plasma particle energy spectra

    International Nuclear Information System (INIS)

    Qin Yunwen

    2001-01-01

    Fitting a curve for an experimental energy spectrum of plasma particles with the least square method, a square error sum between the fitted curve and experimental spectrum data is usually minimized. However, if a straight line is fitted for logarithmic spectrum data to give the particle temperature, it is better to minimize the temperature square error sum. Formulas of such fitting are given and results obtained by minimizing different square error sums are compared

  15. Time and Energy Characterization of a Neutron Time of Flight Detector Using a Novel Coincidence Method for Constraining Neutron Yield, Ion Temperature and Liner Density Measurements from MagLIF Experiments

    Science.gov (United States)

    Styron, Jedediah D.

    The focus of this work is the characterization of a typical neutron time-of-flight (NTOF) detector that is fielded on inertial confinement fusion (ICF) experiments conducted at the Z-experimental facility with emphasis on the Magnetized Liner Fusion (MagLIF) concept. An NTOF detector consisting of EJ-228 scintillator and two independent photomultiplier tubes (PMTs), a Hamamatsu-mod 5 and Photek-PMT240, has been characterized in terms of the absolute time and energy response. The characterization was done by measuring single, neutron-induced events in the scintillator by measuring the alpha particle and neutron produced from the D-T reaction in kinematic coincidence. The results of these experiments provided the time dependent instrument response function and the detector sensitivity as a function of applied voltage covering the entire dynamic range of the detector. Historically, impulse response functions have been measured using various photon sources as surrogates for a neutron source. Thus, this measurement using a single hit neutron source results in the most accurate measured response function, which will improve the accuracy of impulse response corrections to the NTOF signals. While this detector has not yet been fielded on any MagLIF experiments, the development of a predictive capability was desired for transferring the measured detector response from the calibration geometry to the more complex Z geometry. As a proof-of-principle, a detailed model of the Z-machine was developed in MCNP6 to correct for geometry issues when transferring the calibration results from a light lab setting to the Z-environment. Representative values for the instrument response function and the sensitivity for the current detectors fielded on MagLIF experiments were convolved with the modeled results. These results were compared with data collected on three previous MagLIF experiments. The comparison shows the model results can be used to constrain three parameters that are

  16. Track detection methods of radium measurements

    International Nuclear Information System (INIS)

    Somogyi, G.

    1986-06-01

    The principles of tack formation and processing including the description of etching and etch-track evaluation for the preferably used plastic track detectors are discussed. Measuring methods to determine 226 Ra activity based either on the mapping of alpha-decaying elements in the complete U-Ra series by alpha-radiography, or on the measurement of uranium alone by neutron induced fissionography, or on the alpha-decay measurement of 222 Rn, the first daughter element of radium, and finally on the measurement of alpha-tracks originating from radium itself, which is separated from its parent nuclides are described in detail. (V.N.)

  17. Measurement error models, methods, and applications

    CERN Document Server

    Buonaccorsi, John P

    2010-01-01

    Over the last 20 years, comprehensive strategies for treating measurement error in complex models and accounting for the use of extra data to estimate measurement error parameters have emerged. Focusing on both established and novel approaches, ""Measurement Error: Models, Methods, and Applications"" provides an overview of the main techniques and illustrates their application in various models. It describes the impacts of measurement errors on naive analyses that ignore them and presents ways to correct for them across a variety of statistical models, from simple one-sample problems to regres

  18. Composition-independent mean temperature measurements in laminar diffusion flames using spectral lineshape information

    Science.gov (United States)

    Zelenak, D.; Narayanaswamy, V.

    2017-10-01

    Temperature is an important thermochemical property in combusting flows that holds the key to uncovering pollutant formation, flame extinction, and heat release. In a practical combustion environment, the local composition is typically unknown, which hinders the effectiveness of many traditional non-intrusive thermometry techniques. This study aims to offset this limitation by developing a laser-based thermometry technique that does not require prior knowledge of the local composition. Two methods for obtaining temperature are demonstrated in this work, both of which make use of the spectral line broadening of an absorbing species (krypton) seeded into the flow. In the first method, the local Doppler broadening is extracted from an excitation scan to yield the corresponding temperature, while the second method utilizes compositional scaling information of the collisional broadening and collisional shift to determine the temperature. Both methods are demonstrated by measuring the radial temperature profile of a steady laminar CH4/N2 diffusion flame with an air co-flow. The accuracy of the temperature measurements obtained using both methods are evaluated using corresponding temperature profiles determined from computational simulations.

  19. Directional radiometric measurements of row-crop temperatures

    Science.gov (United States)

    Kimes, D. S.; Kirchner, J. A.

    1983-01-01

    The variability of directional sensor response for a cotton row crop in Phoenix, Arizona was measured for various solar zenith angles. The geometric structure of the canopy was described with regard to height, width, spacing, and shape of rows. In addition, radiometric temperature data were collected on four scene components: sunlit and shaded vegetation and sunlit and shaded soil. These data were used to test the predictions and assumptions of a modified version of the row crop model of Jackson et al. (1979), which predicts the thermal infrared response of a sensor as a function of sensor view angle, component temperature, and geometrical structure of the canopy. The field data showed sensor response differentials as great as 16.2 C when going from a zenith view angle of 0 deg to one of 80 deg normal to the row direction. The rms deviation between the predicted and measured sensor response for all measurement periods and view angles was 0.96 C.

  20. Wall surface temperature measurement for the in-pile radioisotope target capsule

    International Nuclear Information System (INIS)

    Liu Yan

    2008-01-01

    The surface temperatures of the radioisotope target capsule is one of the important parameter to its structure integrity. In order to verify the results of theoretical calculation for design, an in-pile surface temperature of capsule wall was measured in Heavy Water Research Reactor (HWRR). The results showed that the temperature were accordant with the theoretical calculation.The measuring methods of the external and internal surface of capsule wall were described also, including instruments selection, the technology of embedded thermocouples and error analysis. (authors)

  1. Sub-Doppler magneto-optical trap temperatures measured using Rydberg tagging

    International Nuclear Information System (INIS)

    Tallant, J.; Overstreet, K. R.; Schwettmann, A.; Shaffer, J. P.

    2006-01-01

    We present measurements of polarization-gradient cooling in a Cs magneto-optic trap (MOT). The temperature in the MOT is found to vary linearly with the light shift. The slope gives a value of C σ =0.25±0.05 while the minimum temperature T 0 =3.8±3.3 μK, in agreement with measurements done in the absence of a magnetic field. The temperature is determined by using a nondestructive method which uses Rydberg time-of-flight velocity distributions of trapped atoms

  2. A Mueller bridge set for cryogenic temperature measurements

    DEFF Research Database (Denmark)

    Diamond, J.M.

    1966-01-01

    An a.c. Mueller bridge set for resistance thermometry at cryogenic temperature is described. A commercial tuned null detector is used at an operating frequency of 1025 c/s. The set includes a high stability oscillator, line reject filter, phase shifter, Q multiplier and selector box. The latter...... permits the dissipation in the thermometers not being measured to be maintained at the operating level with direct current. A temperature change of the order of 10 μdegK can be detected with 10-8 W applied to the thermometer....

  3. A sensitive optical pyrometer for shock-temperature measurements

    Science.gov (United States)

    Boslough, M. B.; Ahrens, T. J.

    1984-01-01

    A new optical system was used to determine temperatures above 2400 K in shocked materials by measuring the spectral radiance of sub-microsecond pulses of light emitted from initially transparent solid samples in the visible and near infrared (450 to 900 nm). The high sensitivity of this optical pyrometer is attributed to the small number of channels, large aperture (0.03 steradian), the large bandwidth per channel (40 nm), and large photodiode detection area (0.2 sq cm). Improved calibration techniques reduce systematic errors encountered in previous shock-temperature experiments.

  4. Direct Measurements of Infrared Normal Spectral Emissivity of Solid Materials for High-Temperature Applications

    Science.gov (United States)

    Hatzl, S.; Kirschner, M.; Lippig, V.; Sander, T.; Mundt, Ch.; Pfitzner, M.

    2013-11-01

    A new facility for the measurement of the normal spectral emissivity of solid materials for high-temperature applications in the thermal steady state was developed at the Bundeswehr University of Munich. The measurements are performed under atmospheric conditions. The facility covers the temperature range between and and wavelengths between and . The principle of operation involves the spectral comparison of a test sample with a reference blackbody and the sample surface temperature determination with a numerical spectral ratio calculation. The optical characteristics of the blackbody and the sample surface temperature determination are discussed in detail. Furthermore, measurement results of the quasi-reference material silicon-carbide under steady-state conditions are presented to validate the measurement method.

  5. Temperature and velocity measurement fields of fluids using a schlieren system.

    Science.gov (United States)

    Martínez-González, Adrian; Guerrero-Viramontes, J A; Moreno-Hernández, David

    2012-06-01

    This paper proposes a combined method for two-dimensional temperature and velocity measurements in liquid and gas flow using a schlieren system. Temperature measurements are made by relating the intensity level of each pixel in a schlieren image to the corresponding knife-edge position measured at the exit focal plane of the schlieren system. The same schlieren images were also used to measure the velocity of the fluid flow. The measurement is made by using particle image velocimetry (PIV). The PIV software used in this work analyzes motion between consecutive schlieren frames to obtain velocity fields. The proposed technique was applied to measure the temperature and velocity fields in the natural convection of water provoked by a heated rectangular plate.

  6. Human performance assessment: methods and measures

    International Nuclear Information System (INIS)

    Andresen, Gisle; Droeivoldsmo, Asgeir

    2000-10-01

    The Human Error Analysis Project (HEAP) was initiated in 1994. The aim of the project was to acquire insights on how and why cognitive errors occur when operators are engaged in problem solving in advanced integrated control rooms. Since human error had not been studied in the HAlden Man-Machine LABoratory (HAMMLAB) before, it was also necessary to carry out research in methodology. In retrospect, it is clear that much of the methodological work is relevant to human-machine research in general, and not only to research on human error. The purpose of this report is, therefore, to give practitioners and researchers an overview of the methodological parts of HEAP. The scope of the report is limited to methods used throughout the data acquisition process, i.e., data-collection methods, data-refinement methods, and measurement methods. The data-collection methods include various types of verbal protocols, simulator logs, questionnaires, and interviews. Data-refinement methods involve different applications of the Eyecon system, a flexible data-refinement tool, and small computer programs used for rearranging, reformatting, and aggregating raw-data. Measurement methods involve assessment of diagnostic behaviour, erroneous actions, complexity, task/system performance, situation awareness, and workload. The report concludes that the data-collection methods are generally both reliable and efficient. The data-refinement methods, however, should be easier to use in order to facilitate explorative analyses. Although the series of experiments provided an opportunity for measurement validation, there are still uncertainties connected to several measures, due to their reliability still being unknown. (Author). 58 refs.,7 tabs

  7. Non-invasicve measurement of the temperature using scanning small coils

    International Nuclear Information System (INIS)

    Ando, Satoshi; Hasegawa, Takeo; Noguchi, Kenichi

    2002-01-01

    The temperature measurement of the tumor is important for hyperthermia. A thermocouple thermometer is stabbed in the living body, and temperature measurement is being done. The problem of this method is that it can get only a part of temperature and it has large invasion to the patient. Therefore, it can't know even if a part of temperature rises. We researched the matter whether temperature measurement could be done without stabbing a thermometer by using the small coils. A magnetic field is born when it is charged with electricity of the high frequency to the objects. It is caught with the coil which put this magnetic field on making it stand opposite to each other. The electric fields decrease with 1/ (2π r). Then, we can calculate the profile of the electric fields using this equation. There are high collation between RF current and temperature in this object. Therefore, we can estimate the temperature distribution in the heating object non-invasively. The results of this study confirmed, there was high correlation between temperature distribution using thermography and distribution of RF currents from our methods

  8. Initial Neutral Particle Analyzer Measurements of Ion Temperature in NSTX

    Energy Technology Data Exchange (ETDEWEB)

    S.S. Medley; R.E. Bell; M.P. Petrov; A.L. Roquemore; and E.V. Suvorkin

    2002-07-08

    The Neutral Particle Analyzer (NPA) diagnostic on the National Spherical Torus Experiment (NSTX) utilizes a PPPL-designed E||B spectrometer which measures the energy spectra of minority H and bulk D species simultaneously with 39 energy channels per mass specie and a time resolution of 1 msec. The calibrated energy range is E = 0.5-150 keV and the energy resolution varies from (delta)E/E = 3-7% over the surface of the microchannel plate detector. The NPA measures thermal Maxwellian ion spectra to obtain line integrated ion temperatures, T{sub i}. For line integral electron densities below neL {approx} 3.5 x 10{sup 19} m{sup -2}, good agreement is observed between the line integrated NPA T{sub i} and the central T{sub i}(0) measured by the spatially localized CHarge Exchange Recombination Spectroscopy (CHERS) diagnostic. However, with increasingly higher n{sub eL} the NPA T{sub i} falls below the central T{sub i}(0) measured by CHERS because the charge exchange neutral emissivity weights the line integrated NPA measurement outboard of the plasma core. An analytic neutral analysis code, DOUBLE, has been applied to the NPA data to correct for this effect and restore agreement with T{sub i}(0) measured by CHERS. A description of the NPA diagnostic on NSTX and initial ion temperature measurements along with an illustration of application of the DOUBLE code are presented.

  9. Articles for high temperature service and methods for their manufacture

    Science.gov (United States)

    Sarrafi-Nour, Reza; Meschter, Peter Joel; Johnson, Curtis Alan; Luthra, Krishan Lal; Rosenzweig, Larry Steven

    2016-06-14

    An article for use in aggressive environments is presented. In one embodiment, the article comprises a substrate and a self-sealing and substantially hermetic sealing layer comprising an alkaline-earth aluminosilicate disposed over the bondcoat. The substrate may be any high-temperature material, including, for instance, silicon-bearing ceramics and ceramic matrix composites. A method for making such articles is also presented. The method comprises providing a substrate; disposing a self-sealing alkaline-earth aluminosilicate layer over the substrate; and heating the sealing layer to a sealing temperature at which at least a portion of the sealing layer will flow.

  10. On the Methods to Measure Powder Flow.

    Science.gov (United States)

    Tan, Geoffrey; Morton, David A V; Larson, Ian

    2015-01-01

    The flow of powders can often play a critical role in the manufacturing of pharmaceutical products. Many of these processes require good, consistent and predictable flow of powders to ensure continuous production of pharmaceutical dosages and to ensure their quality. Therefore, the flow of powders is of paramount importance to the pharmaceutical industry and thus the measuring and evaluating of powder flow is of utmost importance. At present, there are numerous methods in which the flow of powders can be measured. However, due to the complex and environment-dependent nature of powders, no one method exists that is capable of providing a complete picture of the behaviour of powders under dynamic conditions. Some of the most commonly applied methods to measure the flow of powders include: density indices, such as the Carr index and Hausner ratio, powder avalanching, the angle of repose (AOR), flow through an orifice, powder rheometry and shear cell testing.

  11. Generalized temperature measurement equations for Rhodamine B dye solution and its application to microfluidics.

    Science.gov (United States)

    Shah, Jayna J; Gaitan, Michael; Geist, Jon

    2009-10-01

    Temperature mapping based on fluorescent signal intensity ratios is a widely used noncontact approach for investigating temperature distributions in various systems. This noninvasive method is especially useful for applications, such as microfluidics, where accurate temperature measurements are difficult with conventional physical probes. However, the application of a calibration equation to relate fluorescence intensity ratio to temperature is not straightforward when the reference temperature in a given application is different than the one used to derive the calibration equation. In this report, we develop and validate generalized calibration equations that can be applied for any value of reference temperature. Our analysis shows that a simple linear correction for a 40 degrees C reference temperature produces errors in measured temperatures between -3 to 8 degrees C for three previously published sets of cubic calibration equations. On the other hand, corrections based on an exact solution of these equations restrict the errors to those inherent in the calibration equations. The methods described here are demonstrated for cubic calibration equations derived by three different groups, but the general method can be applied to other dyes and calibration equations.

  12. Power System State Estimation Accuracy Enhancement Using Temperature Measurements of Overhead Line Conductors

    Directory of Open Access Journals (Sweden)

    Wydra Michał

    2016-06-01

    Full Text Available Power system state estimation is a process of real-time online modeling of an electric power system. The estimation is performed with the application of a static model of the system and current measurements of electrical quantities that are encumbered with an error. Usually, a model of the estimated system is also encumbered with an uncertainty, especially power line resistances that depend on the temperature of conductors. At present, a considerable development of technologies for dynamic power line rating can be observed. Typically, devices for dynamic line rating are installed directly on the conductors and measure basic electric parameters such as the current and voltage as well as non-electric ones as the surface temperature of conductors, their expansion, stress or the conductor sag angle relative to the plumb line. The objective of this paper is to present a method for power system state estimation that uses temperature measurements of overhead line conductors as supplementary measurements that enhance the model quality and thereby the estimation accuracy. Power system state estimation is presented together with a method of using the temperature measurements of power line conductors for updating the static power system model in the state estimation process. The results obtained with that method have been analyzed based on the estimation calculations performed for an example system - with and without taking into account the conductor temperature measurements. The final part of the article includes conclusions and suggestions for the further research.

  13. Use of aluminum nitride to obtain temperature measurements in a high temperature and high radiation environment

    Science.gov (United States)

    Wernsman, Bernard R.; Blasi, Raymond J.; Tittman, Bernhard R.; Parks, David A.

    2016-04-26

    An aluminum nitride piezoelectric ultrasonic transducer successfully operates at temperatures of up to 1000.degree. C. and fast (>1 MeV) neutron fluencies of more than 10.sup.18 n/cm.sup.2. The transducer comprises a transparent, nitrogen rich aluminum nitride (AlN) crystal wafer that is coupled to an aluminum cylinder for pulse-echo measurements. The transducer has the capability to measure in situ gamma heating within the core of a nuclear reactor.

  14. TWO METHODS FOR REMOTE ESTIMATION OF COMPLETE URBAN SURFACE TEMPERATURE

    Directory of Open Access Journals (Sweden)

    L. Jiang

    2017-09-01

    Full Text Available Complete urban surface temperature (TC is a key parameter for evaluating the energy exchange between the urban surface and atmosphere. At the present stage, the estimation of TC still needs detailed 3D structure information of the urban surface, however, it is often difficult to obtain the geometric structure and composition of the corresponding temperature of urban surface, so that there is still lack of concise and efficient method for estimating the TC by remote sensing. Based on the four typical urban surface scale models, combined with the Envi-met model, thermal radiant directionality forward modeling and kernel model, we analyzed a complete day and night cycle hourly component temperature and radiation temperature in each direction of two seasons of summer and winter, and calculated hemispherical integral temperature and TC. The conclusion is obtained by examining the relationship of directional radiation temperature, hemispherical integral temperature and TC: (1 There is an optimal angle of radiation temperature approaching the TC in a single observation direction when viewing zenith angle is 45–60°, the viewing azimuth near the vertical surface of the sun main plane, the average absolute difference is about 1.1 K in the daytime. (2 There are several (3–5 times directional temperatures of different view angle, under the situation of using the thermal radiation directionality kernel model can more accurately calculate the hemispherical integral temperature close to TC, the mean absolute error is about 1.0 K in the daytime. This study proposed simple and effective strategies for estimating TC by remote sensing, which are expected to improve the quantitative level of remote sensing of urban thermal environment.

  15. Two Methods for Remote Estimation of Complete Urban Surface Temperature

    Science.gov (United States)

    Jiang, L.; Zhan, W.; Zou, Z.

    2017-09-01

    Complete urban surface temperature (TC) is a key parameter for evaluating the energy exchange between the urban surface and atmosphere. At the present stage, the estimation of TC still needs detailed 3D structure information of the urban surface, however, it is often difficult to obtain the geometric structure and composition of the corresponding temperature of urban surface, so that there is still lack of concise and efficient method for estimating the TC by remote sensing. Based on the four typical urban surface scale models, combined with the Envi-met model, thermal radiant directionality forward modeling and kernel model, we analyzed a complete day and night cycle hourly component temperature and radiation temperature in each direction of two seasons of summer and winter, and calculated hemispherical integral temperature and TC. The conclusion is obtained by examining the relationship of directional radiation temperature, hemispherical integral temperature and TC: (1) There is an optimal angle of radiation temperature approaching the TC in a single observation direction when viewing zenith angle is 45-60°, the viewing azimuth near the vertical surface of the sun main plane, the average absolute difference is about 1.1 K in the daytime. (2) There are several (3-5 times) directional temperatures of different view angle, under the situation of using the thermal radiation directionality kernel model can more accurately calculate the hemispherical integral temperature close to TC, the mean absolute error is about 1.0 K in the daytime. This study proposed simple and effective strategies for estimating TC by remote sensing, which are expected to improve the quantitative level of remote sensing of urban thermal environment.

  16. EXPERIMENTAL MEASUREMENT, ANALYSIS AND MODELLING OF DEPENDENCY EMISSIVITY IN FUNCTION OF TEMPERATURE

    Directory of Open Access Journals (Sweden)

    N. Baba Ahmed

    2015-08-01

    Full Text Available We propose a direct method of measurement of the total emissivity of opaque samples on a range of temperature around the ambient one. The method rests on the modulation of the temperature of the sample and the infra-red signal processing resulting from the surface of the sample we model the total emissivity obtained in experiments according to the temperature to establish linear correlations. This leads us to apply the method of optimal linearization associated the finite element method with the nonlinear problem of transfer of heat if thermal conductivity, the specific heat and the emissivity of studied material depend on the temperature. We obtain a good agreement between the resolution of the nonlinear equation of heat and the results obtained by the experimentation. .

  17. Modified methods of stellar magnetic field measurements

    Science.gov (United States)

    Kholtygin, A. F.

    2014-12-01

    The standard methods of the magnetic field measurement, based on an analysis of the relation between the Stokes V-parameter and the first derivative of the total line profile intensity, were modified by applying a linear integral operator \\hat{L} to both sides of this relation. As the operator \\hat{L}, the operator of the wavelet transform with DOG-wavelets is used. The key advantage of the proposed method is an effective suppression of the noise contribution to the line profile and the Stokes parameter V. The efficiency of the method has been studied using model line profiles with various noise contributions. To test the proposed method, the spectropolarimetric observations of the A0 star α2 CVn, the Of?p star HD 148937, and the A0 supergiant HD 92207 were used. The longitudinal magnetic field strengths calculated by our method appeared to be in good agreement with those determined by other methods.

  18. Methods and Devices used to Measure Friction

    DEFF Research Database (Denmark)

    Jeswiet, Jack; Arentoft, Mogens; Henningsen, Poul

    2004-01-01

    . To gain a good understanding of the mechanisms at the interface and to be able to verify the friction and tribology models that exist, friction sensors are needed. Designing sensors to measure friction-stress in metal working has been pursued by many researchers. This paper surveys methods, which have...... been tried in the past and discusses some of the recent sensor designs, which can now be used to measure Friction in both production situations and for research purposes....

  19. Contact Angle Measurement in Lattice Boltzmann Method

    OpenAIRE

    Wen, Binghai; Huang, Bingfang; Qin, Zhangrong; Wang, Chunlei; Zhang, Chaoying

    2017-01-01

    Contact angle is an essential characteristic in wetting, capillarity and moving contact line; however, although contact angle phenomena are effectively simulated, an accurate and real-time measurement for contact angle has not been well studied in computational fluid dynamics, especially in dynamic environments. Here, we design a geometry-based mesoscopic scheme to onthesport measure the contact angle in the lattice Boltzmann method. The computational results without gravity effect are in exc...

  20. Temperature Measurement Technique Using Phosphorescence of Porphyrin Dyes

    Science.gov (United States)

    Kura, Kentaro; Someya, Satoshi; Okamoto, Koji

    2007-11-01

    LIF have been developed to measure the temperature, pH and the oxygen concentration in the fluid. However, the frequent excitation of the fluorescent dye causes the quenching effect. In addition, two color LIF should be applied in order to cancel the effect of non-uniform light intensity of excitation. The phosphor emitting the phosphorescence for a few milliseconds by an excitation was measured at the high time resolution, while the phosphorescence lifetime is the function of the temperature. As the phosphorescence dyes, PtTFPP and PdTFPP were tested. Those mixed with Coumarin30 were also demonstrated. These dyes were excited by a CW laser with the wavelength of 405nm. As the result, it was clarified to be able to measure the temperature using these dyes and this laser. Present study is the result of ``High speed three-dimensional direct measurement technology development for the evaluation of heat flux and flow of liquid metal'' entrusted to the University of Tokyo by the Ministry of Education, Culture, Sports, Science and Technology of Japan (MEXT).

  1. Specific-heat measurement of single metallic, carbon, and ceramic fibers at very high temperature

    International Nuclear Information System (INIS)

    Pradere, C.; Goyheneche, J.M.; Batsale, J.C.; Dilhaire, S.; Pailler, R.

    2005-01-01

    The main objective of this work is to present a method for measuring the specific heat of single metallic, carbon, and ceramic fibers at very high temperature. The difficulty of the measurement is due to the microscale of the fiber (≅10 μm) and the important range of temperature (700-2700 K). An experimental device, a modelization of the thermal behavior, and an analytic model have been developed. A discussion on the measurement accuracy yields a global uncertainty lower than 10%. The characterization of a tungsten filament with thermal properties identical to those of the bulk allows the validation of the device and the thermal estimation method. Finally, measurements on carbon and ceramic fibers have been done at very high temperature

  2. Statistical methods towards more efficient infiltration measurements.

    Science.gov (United States)

    Franz, T; Krebs, P

    2006-01-01

    A comprehensive knowledge about the infiltration situation in a catchment is required for operation and maintenance. Due to the high expenditures, an optimisation of necessary measurement campaigns is essential. Methods based on multivariate statistics were developed to improve the information yield of measurements by identifying appropriate gauge locations. The methods have a high degree of freedom against data needs. They were successfully tested on real and artificial data. For suitable catchments, it is estimated that the optimisation potential amounts up to 30% accuracy improvement compared to nonoptimised gauge distributions. Beside this, a correlation between independent reach parameters and dependent infiltration rates could be identified, which is not dominated by the groundwater head.

  3. Three methods to measure RH bond energies

    Energy Technology Data Exchange (ETDEWEB)

    Berkowitz, J. [Argonne National Lab., IL (United States); Ellison, G.B. [Univ. of Colorado, Boulder, CO (United States). Dept. of Chemistry and Biochemistry; Gutman, D. [Catholic Univ. of America, Washington, DC (United States). Dept. of Chemistry

    1993-03-21

    In this paper the authors compare and contrast three powerful methods for experimentally measuring bond energies in polyatomic molecules. The methods are: radical kinetics; gas phase acidity cycles; and photoionization mass spectroscopy. The knowledge of the values of bond energies are a basic piece of information to a chemist. Chemical reactions involve the making and breaking of chemical bonds. It has been shown that comparable bonds in polyatomic molecules, compared to the same bonds in radicals, can be significantly different. These bond energies can be measured in terms of bond dissociation energies.

  4. Three methods to measure RH bond energies

    International Nuclear Information System (INIS)

    Berkowitz, J.; Ellison, G.B.; Gutman, D.

    1993-01-01

    In this paper the authors compare and contrast three powerful methods for experimentally measuring bond energies in polyatomic molecules. The methods are: radical kinetics; gas phase acidity cycles; and photoionization mass spectroscopy. The knowledge of the values of bond energies are a basic piece of information to a chemist. Chemical reactions involve the making and breaking of chemical bonds. It has been shown that comparable bonds in polyatomic molecules, compared to the same bonds in radicals, can be significantly different. These bond energies can be measured in terms of bond dissociation energies

  5. Magnetic hyperthermia heating of cobalt ferrite nanoparticles prepared by low temperature ferrous sulfate based method

    Directory of Open Access Journals (Sweden)

    Tejabhiram Yadavalli

    2016-05-01

    Full Text Available A facile low temperature co-precipitation method for the synthesis of crystalline cobalt ferrite nanostructures using ferrous sulfate salt as the precursor has been discussed. The prepared samples were compared with nanoparticles prepared by conventional co-precipitation and hydrothermal methods using ferric nitrate as the precursor. X-ray diffraction studies confirmed the formation of cubic spinel cobalt ferrites when dried at 110 °C as opposed to conventional methods which required higher temperatures/pressure for the formation of the same. Field emission scanning electron microscope studies of these powders revealed the formation of nearly spherical nanostructures in the size range of 20-30 nm which were comparable to those prepared by conventional methods. Magnetic measurements confirmed the ferromagnetic nature of the cobalt ferrites with low magnetic remanance. Further magnetic hyperthermia studies of nanostructures prepared by low temperature method showed a rise in temperature to 50 °C in 600 s.

  6. A new method to measure vaginal sensibility.

    Science.gov (United States)

    Lakeman, M M E; Laan, E; Vaart, C H; Roovers, J P

    2010-06-01

    Vaginal surgery may affect sexual function both positively and negatively. Possibly, negative consequences of surgical interventions on sexuality may be caused by reduced sensibility of the vaginal wall. To develop a new method to measure vaginal sensibility. We developed a technique to measure the sensibility of the vaginal wall consisting of a St Marks electrode on a gloved index finger, with a stimulating electrode mounted at the tip. Measurements were performed in four different target areas (caudal and cranial, posterior and anterior) by two independent female researchers in a random order. Subjects were 12 healthy women. The intra-observer reproducibility of both researchers was almost perfect (Pearsons-Rho correlation coefficient 0.77-0.96 sensibility in the cranial posterior vaginal wall relative to the cranial anterior vaginal wall, but for all measurements, researcher 2 obtained higher sensibility ratings than researcher 1. In addition, researcher 2 found a decreased sensibility in the cranial anterior vaginal wall for women not using oral contraceptives. Phase of the menstrual cycle did not influence vaginal sensibility. We developed a new instrument to measure vaginal sensibility. The instrument has excellent intra-observer reproducibility. This method is sufficiently sensitive so as to differentiate between anterior and posterior cranial vaginal wall sensibility, but outcome differs as a function of researcher. Further evaluation of the clinical use of this method is needed, provided that measurements are performed by the same researcher. (c) 2010 Wiley-Liss, Inc.

  7. Thermocouple and infrared sensor-based measurement of temperature distribution in metal cutting.

    Science.gov (United States)

    Kus, Abdil; Isik, Yahya; Cakir, M Cemal; Coşkun, Salih; Özdemir, Kadir

    2015-01-12

    In metal cutting, the magnitude of the temperature at the tool-chip interface is a function of the cutting parameters. This temperature directly affects production; therefore, increased research on the role of cutting temperatures can lead to improved machining operations. In this study, tool temperature was estimated by simultaneous temperature measurement employing both a K-type thermocouple and an infrared radiation (IR) pyrometer to measure the tool-chip interface temperature. Due to the complexity of the machining processes, the integration of different measuring techniques was necessary in order to obtain consistent temperature data. The thermal analysis results were compared via the ANSYS finite element method. Experiments were carried out in dry machining using workpiece material of AISI 4140 alloy steel that was heat treated by an induction process to a hardness of 50 HRC. A PVD TiAlN-TiN-coated WNVG 080404-IC907 carbide insert was used during the turning process. The results showed that with increasing cutting speed, feed rate and depth of cut, the tool temperature increased; the cutting speed was found to be the most effective parameter in assessing the temperature rise. The heat distribution of the cutting tool, tool-chip interface and workpiece provided effective and useful data for the optimization of selected cutting parameters during orthogonal machining.

  8. Thermocouple and Infrared Sensor-Based Measurement of Temperature Distribution in Metal Cutting

    Directory of Open Access Journals (Sweden)

    Abdil Kus

    2015-01-01

    Full Text Available In metal cutting, the magnitude of the temperature at the tool-chip interface is a function of the cutting parameters. This temperature directly affects production; therefore, increased research on the role of cutting temperatures can lead to improved machining operations. In this study, tool temperature was estimated by simultaneous temperature measurement employing both a K-type thermocouple and an infrared radiation (IR pyrometer to measure the tool-chip interface temperature. Due to the complexity of the machining processes, the integration of different measuring techniques was necessary in order to obtain consistent temperature data. The thermal analysis results were compared via the ANSYS finite element method. Experiments were carried out in dry machining using workpiece material of AISI 4140 alloy steel that was heat treated by an induction process to a hardness of 50 HRC. A PVD TiAlN-TiN-coated WNVG 080404-IC907 carbide insert was used during the turning process. The results showed that with increasing cutting speed, feed rate and depth of cut, the tool temperature increased; the cutting speed was found to be the most effective parameter in assessing the temperature rise. The heat distribution of the cutting tool, tool-chip interface and workpiece provided effective and useful data for the optimization of selected cutting parameters during orthogonal machining.

  9. Thermocouple and Infrared Sensor-Based Measurement of Temperature Distribution in Metal Cutting

    Science.gov (United States)

    Kus, Abdil; Isik, Yahya; Cakir, M. Cemal; Coşkun, Salih; Özdemir, Kadir

    2015-01-01

    In metal cutting, the magnitude of the temperature at the tool-chip interface is a function of the cutting parameters. This temperature directly affects production; therefore, increased research on the role of cutting temperatures can lead to improved machining operations. In this study, tool temperature was estimated by simultaneous temperature measurement employing both a K-type thermocouple and an infrared radiation (IR) pyrometer to measure the tool-chip interface temperature. Due to the complexity of the machining processes, the integration of different measuring techniques was necessary in order to obtain consistent temperature data. The thermal analysis results were compared via the ANSYS finite element method. Experiments were carried out in dry machining using workpiece material of AISI 4140 alloy steel that was heat treated by an induction process to a hardness of 50 HRC. A PVD TiAlN-TiN-coated WNVG 080404-IC907 carbide insert was used during the turning process. The results showed that with increasing cutting speed, feed rate and depth of cut, the tool temperature increased; the cutting speed was found to be the most effective parameter in assessing the temperature rise. The heat distribution of the cutting tool, tool-chip interface and workpiece provided effective and useful data for the optimization of selected cutting parameters during orthogonal machining. PMID:25587976

  10. Ratio method of measuring W boson mass

    Energy Technology Data Exchange (ETDEWEB)

    Guo, Feng [Stony Brook Univ., NY (United States)

    2010-08-01

    This dissertation describes an alternative method of measuring the W boson mass in DØ experiment. Instead of extracting MW from the fitting of W → ev fast Monte Carlo simulations to W → ev data as in the standard method, we make the direct fit of transverse mass between W → ev data and Z → ee data. One of the two electrons from Z boson is treated as a neutrino in the calculation of transverse mass. In ratio method, the best fitted scale factor corresponds to the ratio of W and Z boson mass (MW/MZ). Given the precisely measured Z boson mass, W mass is directly fitted from W → ev and Z → ee data. This dissertation demonstrates that ratio method is a plausible method of measuring the W boson mass. With the 1 fb-1 DØ Run IIa dataset, ratio method gives MW = 80435 ± 43(stat) ± 26(sys) MeV.

  11. Measuring core body temperature with a non-invasive sensor.

    Science.gov (United States)

    Mazgaoker, Savyon; Ketko, Itay; Yanovich, Ran; Heled, Yuval; Epstein, Yoram

    2017-05-01

    In various occupations, workers may be exposed to extreme environmental conditions and physical activities. Under these conditions the ability to follow the workers' body temperature may protect them from overheating that may lead to heat related injuries. The "Dräger" Double Sensor (DS) is a novel device for assessing body-core temperature (T c ). The purpose of this study was to evaluate the accuracy of the DS in measuring T c under heat stress. Seventeen male participants performed a three stage protocol: 30min rest in a thermal comfort environment (20-22°C, 50% relative humidity), followed by an exposure to a hot environment of 40°C, 40% relative humidity -30min at rest and 60min of exercise (walking on a treadmill at 5km/h and 2% elevation). Simultaneously temperatures measured by the DS (T DS ) and by rectal temperature (T re ) (YSI-401 thermistor) were recorded and then compared. During the three stages of the study the average temperature obtained by the DS was within±0.3°C of rectal measurement. The correlation between T DS and T re was significantly better during the heat exposures phases than during resting under comfort conditions. These preliminary results are promising for potential use of the DS by workers under field conditions and especially under environmental heat stress or when dressed in protective garments. For this goal, further investigations are required to validate the accuracy of the DS under various levels of heat stress, clothing and working levels. Copyright © 2017. Published by Elsevier Ltd.

  12. Method of superconducting joint and its measurement

    International Nuclear Information System (INIS)

    Kim, Woo Gon; Lee, Ho Jin; Hong, Gye Won

    1994-04-01

    The development of joint techniques for superconducting wires is essential to fabricate the high quality superconducting magnet. In this report, the various joining methods and their measuring techniques were reviewed. In order to fabricate a precise superconducting magnet, joining and measuring experiment by using the field decay technique carried out. The contact resistance of coupled specimens with joint was measured as 3.0 x 10 -15 ohm at 1 Tesla which is lower than that of the real operating condition of MRI magnet. It is expected that these data can be used to design and fabricate the superconducting magnets successfully. (Author) 12 refs., 20 figs., 2 tabs

  13. Comparison Testings between Two High-temperature Strain Measurement Systems

    Science.gov (United States)

    Lei, J.-F.; Castelli, M. G.; Androjna, D.; Blue, C.; Blue, R.; Lin, R. Y.

    1996-01-01

    An experimental evaluation was conducted at NASA Lewis Research Center to compare and contrast the performance of a newly developed resistance strain gage, the PdCr temperature-compensated wire strain gage, to that of a conventional high-temperature extensometry. The evaluation of the two strain measurement systems was conducted through the application of various thermal and mechanical loading spectra using a high-temperature thermomechanical uniaxial testing system equipped with quartz lamp heating. The purpose of the testing was not only to compare and contrast the two strain sensors but also to investigate the applicability of the PdCr strain gage to the testing environment typically employed when characterizing the high-temperature mechanical behavior of structural materials. Strain measurement capabilities to 8OO C were investigated with a nickel base superalloy IN100 substrate material, and application to titanium matrix composite (TMC) materials was examined with the SCS-6/Ti-15-3 08 system. PdCr strain gages installed by three attachment techniques, namely, flame spraying, spot welding and rapid infrared joining were investigated.

  14. Micrometeorological methods for measurements of mercury emissions over contaminated soils

    International Nuclear Information System (INIS)

    Kim, K.H.; Lindberg, S.E.; Hanson, P.J.; Owens, J.; Myers, T.P.

    1993-01-01

    As part of a larger study involving development and application of field and laboratory methods (micrometeorological, dynamic enclosure chamber, and controlled laboratory chamber methods) to measure the air/surface exchange of Hg vapor, we performed a series of preliminary measurements over contaminated soils. From March--April 1993, we used the modified Bowen ratio (MBR) method to measure emission rates of mercury over a floodplain contaminated with mercury near Oak Ridge, TN. The mercury emission rates measured from contaminated EFPC soils using the MBR method during early spring show that (1) in all cases, the contaminated soils acted as a source of mercury to the atmosphere with source strengths ranging from 17 to 160 ng m -2 h -1 ; and (2) the strengths of mercury emissions can be greatly influenced by the combined effects of surface soil temperature, residence time of air masses over the source area, and turbulence conditions. The mercury fluxes measured in a controlled flow chamber indicate that contaminated soils can exhibit up to an order of magnitude higher emission rates of Hg under conditions of elevated soil temperature, soil structure disturbance, and high turbulence. Mercury emissions from contaminated soils exceeded emissions from background soils by one to two orders of magnitude

  15. Using Distributed Temperature Sensing for evaporation measurements: background, verification, and future applications.

    Science.gov (United States)

    Schilperoort, Bart; Coenders-Gerrits, Miriam; van Iersel, Tara; Jiménez Rodríguez, Cesar; Luxemburg, Willem; Cisneros Vaca, Cesar; Ucer, Murat

    2017-04-01

    Distributed temperature sensing (DTS) is a relatively new method for measuring latent and sensible heat fluxes. The method has been successfully tested before on multiple sites (Euser, 2014). It uses a glass fibre optic cable of which the temperature can be measured every 12.5cm. By placing the cable vertically along a structure, the air temperature profile can be measured. If the cable is wrapped with cloth and kept wet (akin to a psychrometer), a vertical wet-bulb temperature gradient over height can be calculated. From these dry and wet-bulb temperatures over the height the Bowen ratio is determined and together with the energy balance the latent and sensible heat can be determined. To verify the measurements of the DTS based Bowen ratio method (BR-DTS) we assessed in detail; the accuracy of the air temperature and wet-bulb temperature measurements, the influence of solar radiation and wind on these temperatures, and a comparison to standard methods of evaporation measurement. We tested the performance of the BR-DTS on a 45m high tower in a tall mixed forest in the centre of the Netherlands in August. The average tree height is 30m, hence we measure temperature gradients above, in, and underneath the canopy. We found that solar radiation has a significant effect on the temperature measurements due to heating of the cable coating and leads to deviations up to 2° C. By using cables with different coating thickness we could theoretically correct for this effect, but this introduces too much uncertainty for calculating the temperature gradient. By installing screens the effect of direct sunlight on the cable is sufficiently reduced, and the correlation of the cable temperature with reference air temperature sensors is very high (R2=0.988 to 0.998). Wind speed seems to have a minimal effect on the measured wet-bulb temperature, both below and above the canopy. The latent heat fluxes of the BR-DTS were compared to an eddy covariance system using data from 10 days

  16. Measurement methods of building structures deflections

    Directory of Open Access Journals (Sweden)

    Wróblewska Magdalena

    2018-01-01

    Full Text Available Underground mining exploitation is leading to the occurrence of deformations manifested by, in particular, sloping terrain. The structures situated on the deforming subsoil are subject to uneven subsidence which is leading in consequence to their deflection. Before a building rectification process takes place by, e.g. uneven raising, the structure's deflection direction and value is determined so that the structure is restored to its vertical position as a result of the undertaken remedial measures. Deflection can be determined by applying classical as well as modern measurement techniques. The article presents examples of measurement methods used considering the measured elements of building structures’ constructions and field measurements. Moreover, for a given example of a mining area, the existing deflections of buildings were compared with mining terrain sloping.

  17. Rock mass and shaft concrete lining temperature measurement procedure: Final draft

    International Nuclear Information System (INIS)

    1986-10-01

    This procedure document describes the equipment and procedures which will be used to obtain temperature data from within rock-mass and shaft linings at the Deaf Smith Exploratory Shaft Facility. Temperature measurement methods for instrument temperature correction, fluid temperature correction, heated surface monitoring and air temperature monitoring are outside the scope of this procedure, and are covered in the appropriate individual test procedures. Calibration, acceptance testing and the assignment of transducer reference numbers are outside the scope of this procedure. Section 2.0 provides a summary of the temperature measurement methods which will be employed, together with the measurement locations, environmental considerations and measurement requirements. Test layouts, including detailed descriptions of instruments, support requirements and detailed installation procedures are also presented. Section 3.0 describes the requirements for data recording, ADAS monitoring, and data reporting. Section 4.0 defines personnel responsibilities and qualifications. In addition a measurement and installation schedule is provided, and safety and contingency plans are described. Section 5.0 discusses management and quality assurance requirements. Cited references are listed in Section 6.0. 7 refs., 9 figs

  18. Measurement of Laser Weld Temperatures for 3D Model Input

    Energy Technology Data Exchange (ETDEWEB)

    Dagel, Daryl [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Grossetete, Grant [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States); Maccallum, Danny O. [Sandia National Lab. (SNL-NM), Albuquerque, NM (United States)

    2016-10-01

    Laser welding is a key joining process used extensively in the manufacture and assembly of critical components for several weapons systems. Sandia National Laboratories advances the understanding of the laser welding process through coupled experimentation and modeling. This report summarizes the experimental portion of the research program, which focused on measuring temperatures and thermal history of laser welds on steel plates. To increase confidence in measurement accuracy, researchers utilized multiple complementary techniques to acquire temperatures during laser welding. This data serves as input to and validation of 3D laser welding models aimed at predicting microstructure and the formation of defects and their impact on weld-joint reliability, a crucial step in rapid prototyping of weapons components.

  19. High temperature elastic constant measurements: application to plutonium

    International Nuclear Information System (INIS)

    Bouchet, J.M.

    1969-03-01

    We present an apparatus with which we have measured the Young's modulus and the Poisson's ratio of several compounds from the resonance frequency of cylinders in the temperature range 0 deg. C-700 deg. C. We especially studied the elastic constants of plutonium and measured for the first time to our knowledge the Young's modulus of Pu δ and Pu ε . E δ 360 deg. C = 1.6 10 11 dy/cm 2 ; E ε 490 deg. C = 1.1 10 11 dy/cm 2 , σ ε = 0.25 ± 0.03 Using our results, we have calculated the compressibility, the Debye temperature, the Grueneisen constant and the electronic specific heat of Pu ε . (author) [fr

  20. Dual purpose pyrometer for temperature and solidification velocity measurement

    Science.gov (United States)

    Hofmeister, W. H.; Bayuzick, R. J.; Robinson, M. B.

    1990-01-01

    A dual purpose pyrometer is described that allows both accurate radiance temperature measurement and fast temporal response. The system uses two silicon photodiodes with separate optical paths derived from a common spot on the sample. The optical bandwidths and response times of each detection circuit are tailored to the function of each radiometer. The radiance temperature of electromagnetically levitated metallic samples is measured over a narrow optical bandwidth with a high-gain silicon detector. The velocity of solidification of undercooled melts can be deduced from the rise time of the second silicon detector which samples a broad optical bandwidth and has a fast response time. Results from experiments on the undercooling and solidification behavior of electromagnetically levitated pure nickel show that the solidification velocity approaches 17 m/s at high undercooling.