WorldWideScience

Sample records for reached alarming levels

  1. Alarming levels of carboxyhemoglobin in banked blood.

    Science.gov (United States)

    Ehlers, Melissa; Labaze, Georges; Hanakova, Marcela; McCloskey, David; Wilner, George

    2009-06-01

    To determine the level of carboxyhemoglobin found in banked blood in the Albany, NY region. A retrospective descriptive analysis of carboxyhemoglobin (COHb) levels in a series of packed red blood cell (PRBC) units. The blood bank of a university tertiary care hospital in Albany, NY. All PRBC units considered for possible use in pediatric cardiac surgery were first analyzed for levels of COHb. Only those units with COHb levels of <1.5% were deemed acceptable for use during pediatric cardiac surgery. A sample of blood drawn from the sample side arm of each PRBC unit was analyzed on a Chiron 855 Blood Gas Analyzer (Chiron Inc, Emeryville, CA, now Siemens/Bayer RapidLab 865) to determine the level of COHb. The average COHb level was 0.78% (standard deviation +/- 1.48%), and out of the 468 units tested, 48 (10.3%) had COHb levels of 1.5% or greater. The highest recorded COHb level was 12%. The transfusion of PRBC units may artificially elevate readings of COHb and cause confusion over possible causes. Certain high-risk populations (eg, cyanotic neonates undergoing cardiopulmonary bypass) may be especially at risk. Although levels of COHb in the US blood supply are dropping, institutions may want to consider analyzing COHb levels in their PRBC units before transfusion in these high-risk populations.

  2. Levels of alarm thresholds of meningitis outbreaks in Hamadan Province, west of Iran.

    Science.gov (United States)

    Faryadres, Mohammad; Karami, Manoochehr; Moghimbeigi, Abbas; Esmailnasab, Nader; Pazhouhi, Khabat

    2015-01-01

    Few studies have focused on syndromic data to determine levels of alarm thresholds to detection of meningitis outbreaks. The purpose of this study was to determine threshold levels of meningitis outbreak in Hamadan Province, west of Iran. Data on both confirmed and suspected cases of meningitis (fever and neurological symptom) form 21 March 2010 to 20 March 2012 were used in Hamadan Province, Iran. Alarm threshold levels of meningitis outbreak were determined using four different methods including absolute values or standard method, relative increase, statistical cutoff points and upper control limit of exponentially weighted moving average (EWMA) algorithm. Among 723 reported cases, 41 were diagnosed to have meningitis. Standard level of alarm thresholds for meningitis outbreak was determined as incidence of 5/100000 persons. Increasing 1.5 to two times in reported cases of suspected meningitis per week was known as the threshold levels according to relative increase method. An occurrence four cases of suspected meningitis per week that equals to 90th percentile was chosen as alarm thresholds by statistical cut off point method. The corresponding value according to EWMA algorithm was 2.57 i.e. three cases. Policy makers and staff of syndromic surveillance systems are highly recommended to apply the above different methods to determine the levels of alarm threshold.

  3. The challenge of localizing vehicle backup alarms: Effects of passive and electronic hearing protectors, ambient noise level, and backup alarm spectral content

    Directory of Open Access Journals (Sweden)

    Khaled A Alali

    2011-01-01

    Full Text Available A human factors experiment employed a hemi-anechoic sound field in which listeners were required to localize a vehicular backup alarm warning signal (both a standard and a frequency-augmented alarm in 360-degrees azimuth in pink noise of 60 dBA and 90 dBA. Measures of localization performance included: (1 percentage correct localization, (2 percentage of right--left localization errors, (3 percentage of front-rear localization errors, and (4 localization absolute deviation in degrees from the alarm′s actual location. In summary, the data demonstrated that, with some exceptions, normal hearing listeners′ ability to localize the backup alarm in 360-degrees azimuth did not improve when wearing augmented hearing protectors (including dichotic sound transmission earmuffs, flat attenuation earplugs, and level-dependent earplugs as compared to when wearing conventional passive earmuffs or earplugs of the foam or flanged types. Exceptions were that in the 90 dBA pink noise, the flat attenuation earplug yielded significantly better accuracy than the polyurethane foam earplug and both the dichotic and the custom-made diotic electronic sound transmission earmuffs. However, the flat attenuation earplug showed no benefit over the standard pre-molded earplug, the arc earplug, and the passive earmuff. Confusions of front-rear alarm directions were most significant in the 90 dBA noise condition, wherein two types of triple-flanged earplugs exhibited significantly fewer front-rear confusions than either of the electronic muffs. On all measures, the diotic sound transmission earmuff resulted in the poorest localization of any of the protectors due to the fact that its single-microphone design did not enable interaural cues to be heard. Localization was consistently more degraded in the 90 dBA pink noise as compared with the relatively quiet condition of the 60 dBA pink noise. A frequency-augmented backup alarm, which incorporated 400 Hz and 4000 Hz components

  4. Optimal Alarm Systems

    Data.gov (United States)

    National Aeronautics and Space Administration — An optimal alarm system is simply an optimal level-crossing predictor that can be designed to elicit the fewest false alarms for a fixed detection probability. It...

  5. Development of an alarm analysis system based on multi-level flow models for nuclear power plant

    International Nuclear Information System (INIS)

    Zhang Jiande; Yang Ming; Zhang Zhijian

    2008-01-01

    An alarm analysis system based on Multi-level Flow Models (MFM) was developed for a PWR NPP. By automatically identifying the primary root causes in complex fault situations, the workload of the operators can be reduced. In addition, because MFM also provides a set of graphical symbols that implies causalities, operators can confirm diagnosis results by semiotic analysis, and hence the understandability of the process of alarm analysis as well as the reliability of maintenance task can be increased. 19 cases of simulation data from RELAP5/MOD2 code were utilized for evaluating the performance of the proposed system. The simulation results show that the proposed alarm analysis system has a good ability to detect and diagnose accidents earlier in time before reactor trip. (authors)

  6. Alarms, Chemical

    Science.gov (United States)

    cited in applicable qualitative materiel requirements, small development requirements, technical characteristics, and other requirements and documentation that pertain to automatic chemical agent alarms.

  7. Alarming atmospheres

    DEFF Research Database (Denmark)

    Højlund, Marie; Kinch, Sofie

    2014-01-01

    Nurses working in the Neuro-Intensive Care Unit at Aarhus University Hospital lack the tools to prepare children for the alarming atmosphere they will enter when visiting a hospitalised relative. The complex soundscape dominated by alarms and sounds from equipment is mentioned as the main stressor...

  8. Alarming levels of drug-resistant tuberculosis in HIV-infected patients in metropolitan Mumbai, India.

    Science.gov (United States)

    Isaakidis, Petros; Das, Mrinalini; Kumar, Ajay M V; Peskett, Christopher; Khetarpal, Minni; Bamne, Arun; Adsul, Balkrishna; Manglani, Mamta; Sachdeva, Kuldeep Singh; Parmar, Malik; Kanchar, Avinash; Rewari, B B; Deshpande, Alaka; Rodrigues, Camilla; Shetty, Anjali; Rebello, Lorraine; Saranchuk, Peter

    2014-01-01

    Drug-resistant tuberculosis (DR-TB) is a looming threat to tuberculosis control in India. However, no countrywide prevalence data are available. The burden of DR-TB in HIV-co-infected patients is likewise unknown. Undiagnosed and untreated DR-TB among HIV-infected patients is a major cause of mortality and morbidity. We aimed to assess the prevalence of DR-TB (defined as resistance to any anti-TB drug) in patients attending public antiretroviral treatment (ART) centers in greater metropolitan Mumbai, India. A cross-sectional survey was conducted among adults and children ART-center attendees. Smear microscopy, culture and drug-susceptibility-testing (DST) against all first and second-line TB-drugs using phenotypic liquid culture (MGIT) were conducted on all presumptive tuberculosis patients. Analyses were performed to determine DR-TB prevalence and resistance patterns separately for new and previously treated, culture-positive TB-cases. Between March 2013 and January 2014, ART-center attendees were screened during 14135 visits, of whom 1724 had presumptive TB. Of 1724 attendees, 72 (4%) were smear-positive and 202 (12%) had a positive culture for Mycobacterium tuberculosis. Overall DR-TB was diagnosed in 68 (34%, 95% CI: 27%-40%) TB-patients. The proportions of DR-TB were 25% (29/114) and 44% (39/88) among new and previously treated cases respectively. The patterns of DR-TB were: 21% mono-resistant, 12% poly-resistant, 38% multidrug-resistant (MDR-TB), 21% pre-extensively-drug-resistant (MDR-TB plus resistance to either a fluoroquinolone or second-line injectable), 6% extensively drug-resistant (XDR-TB) and 2% extremely drug-resistant TB (XDR-TB plus resistance to any group-IV/V drug). Only previous history of TB was significantly associated with the diagnosis of DR-TB in multivariate models. The burden of DR-TB among HIV-infected patients attending public ART-centers in Mumbai was alarmingly high, likely representing ongoing transmission in the community and

  9. Alarming levels of drug-resistant tuberculosis in HIV-infected patients in metropolitan Mumbai, India.

    Directory of Open Access Journals (Sweden)

    Petros Isaakidis

    Full Text Available BACKGROUND: Drug-resistant tuberculosis (DR-TB is a looming threat to tuberculosis control in India. However, no countrywide prevalence data are available. The burden of DR-TB in HIV-co-infected patients is likewise unknown. Undiagnosed and untreated DR-TB among HIV-infected patients is a major cause of mortality and morbidity. We aimed to assess the prevalence of DR-TB (defined as resistance to any anti-TB drug in patients attending public antiretroviral treatment (ART centers in greater metropolitan Mumbai, India. METHODS: A cross-sectional survey was conducted among adults and children ART-center attendees. Smear microscopy, culture and drug-susceptibility-testing (DST against all first and second-line TB-drugs using phenotypic liquid culture (MGIT were conducted on all presumptive tuberculosis patients. Analyses were performed to determine DR-TB prevalence and resistance patterns separately for new and previously treated, culture-positive TB-cases. RESULTS: Between March 2013 and January 2014, ART-center attendees were screened during 14135 visits, of whom 1724 had presumptive TB. Of 1724 attendees, 72 (4% were smear-positive and 202 (12% had a positive culture for Mycobacterium tuberculosis. Overall DR-TB was diagnosed in 68 (34%, 95% CI: 27%-40% TB-patients. The proportions of DR-TB were 25% (29/114 and 44% (39/88 among new and previously treated cases respectively. The patterns of DR-TB were: 21% mono-resistant, 12% poly-resistant, 38% multidrug-resistant (MDR-TB, 21% pre-extensively-drug-resistant (MDR-TB plus resistance to either a fluoroquinolone or second-line injectable, 6% extensively drug-resistant (XDR-TB and 2% extremely drug-resistant TB (XDR-TB plus resistance to any group-IV/V drug. Only previous history of TB was significantly associated with the diagnosis of DR-TB in multivariate models. CONCLUSION: The burden of DR-TB among HIV-infected patients attending public ART-centers in Mumbai was alarmingly high, likely representing

  10. The Level of Europium-154 Contaminating Samarium-153-EDTMP Activates the Radiation Alarm System at the US Homeland Security Checkpoints

    Directory of Open Access Journals (Sweden)

    Mohammed Najeeb Al Hallak

    2009-08-01

    Full Text Available 153Sm-EDTMP is a radiopharmaceutical composed of EDTMP (ethylenediamine-tetramethylenephosphonate and Samarium-153 [1]. 153Sm-EDTMP has an affinity for skeletal tissue and concentrates in areas with increased bone turnover; thus, it is successfully used in relieving pain related to diffuse bone metastases [1]. The manufacturing process of 153Sm-EDTMP leads to contamination with 154Eu (Europium-154 [2]. A previous study only alluded to the retention of 154Eu in the bones after receiving treatment with 153Sm-EDTMP [2]. Activation of the alarm at security checkpoints after 153Sm-EDTMP therapy has not been previously reported. Two out of 15 patients who received 153Sm-EDTMP at Roger Maris Cancer Center (Fargo, N. Dak., USA activated the radiation activity sensors while passing through checkpoints; one at a US airport and the other while crossing theAmerican-Canadian border. We assume that the 154Eu which remained in the patients’ bones activated the sensors. Methods: In order to investigate this hypothesis, we obtained the consent from 3 of our 15 patients who received 153Sm-EDTMP within the previous 4 months to 2 years, including the patient who had activated the radiation alarm at the airport. The patients were scanned with a handheld detector and a gamma camera for energies from 511 keV to 1.3 MeV. Results: All three patients exhibited identical spectral images, and further analysis showed that the observed spectra are the result of 154Eu emissions. Conclusion: Depending on the detection thresholds and windows used by local and federal authorities, the remaining activity of 154Eu retained in patients who received 153Sm-EDTMP could be sufficient enough to increase the count rates above background levels and activate the sensors. At Roger Maris Cancer Center, patients are now informed of the potential consequences of 153Sm-EDTMP therapy prior to initiating treatment. In addition, patients treated with 153Sm-EDTMP at Roger Maris Cancer Center

  11. Criticality alarm device

    International Nuclear Information System (INIS)

    Kasai, Kenji.

    1994-01-01

    The device of the present invention is utilized, for example, to a reprocessing facility for storing and processing nuclear fuels and measures and controls the nuclear fuel assembly system so as not to exceed criticality. That is, a conventional criticality alarm device applies a predetermined processing to neutron fluxes generated from a nuclear fuel assembly system containing nuclear fuels and outputs an alarm. The device of the present invention comprises (1) a neutron flux supply source for increasing and decreasing neutron fluxes periodically and supplying them to nuclear fuel assemblies, (2) a detector for detecting neutron fluxes in the nuclear fuel assemblies, (3) a critical state judging section for judging the critical state of the nuclear fuel assemblies based on the periodically changing signals obtained from the detector (2) and (4) an alarm section for outputting criticality alarms depending on the result of the judgement. The device of the present invention can accurately recognize the critical state of the nuclear fuel assembly system and can forecast reaching of the nuclear fuel assembly to criticality or prompt neutron critical state. (I.S.)

  12. Gynecological cancer alarm symptoms:

    DEFF Research Database (Denmark)

    Balasubramaniam, Kirubakaran; Ravn, Pernille; dePont Christensen, René

    2016-01-01

    INTRODUCTION: To determine the proportion of patients who were referred to specialist care after reporting gynecological cancer alarm symptoms to their general practitioner. To investigate whether contact with specialist care was associated with lifestyle factors or socioeconomic status. MATERIAL...... and odds ratios (ORs) for associations between specialist care contact, lifestyle factors and socioeconomic status. RESULTS: The study included 25 866 non-pregnant women; 2957 reported the onset of at least one gynecological cancer alarm symptom, and 683 of these (23.1%) reported symptoms to their general......: Educational level influence contact with specialist care among patients with gynecological cancer alarm symptoms. Future studies should investigate inequalities in access to the secondary healthcare system. This article is protected by copyright. All rights reserved....

  13. Alarm filtering and presentation

    International Nuclear Information System (INIS)

    Bray, M.A.

    1989-01-01

    This paper discusses alarm filtering and presentation in the control room of nuclear and other process control plants. Alarm generation and presentation is widely recognized as a general process control problem. Alarm systems often fail to provide meaningful alarms to operators. Alarm generation and presentation is an area in which computer aiding is feasible and provides clear benefits. Therefore, researchers have developed several computerized alarm filtering and presentation approaches. This paper discusses problems associated with alarm generation and presentation. Approaches to improving the alarm situation and installation issues of alarm system improvements are discussed. The impact of artificial intelligence (AI) technology on alarm system improvements is assessed. (orig.)

  14. TMACS test procedure TP001: Alarm management. Revision 6

    International Nuclear Information System (INIS)

    Scanlan, P.K.

    1994-01-01

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addresses the Alarm Management requirements of the TMACS. The features to be tested are: real-time alarming on high and low level and discrete alarms, equipment alarms, dead-band filtering, alarm display color coding, alarm acknowledgement, and alarm logging

  15. The Remaining Service Time Upon Reaching a High Level in M/G/1 Queues

    NARCIS (Netherlands)

    de Boer, Pieter-Tjerk; Nicola, V.F.; van Ommeren, Jan C.W.

    The distribution of the remaining service time upon reaching some target level in an M/G/1 queue is of theoretical as well as practical interest. In general, this distribution depends on the initial level as well as on the target level, say, B. Two initial levels are of particular interest, namely,

  16. Ship Detection in Optical Remote Sensing Images Based on Wavelet Transform and Multi-Level False Alarm Identification

    Directory of Open Access Journals (Sweden)

    Fang Xu

    2017-09-01

    Full Text Available Ship detection by Unmanned Airborne Vehicles (UAVs and satellites plays an important role in a spectrum of related military and civil applications. To improve the detection efficiency, accuracy, and speed, a novel ship detection method from coarse to fine is presented. Ship targets are viewed as uncommon regions in the sea background caused by the differences in colors, textures, shapes, or other factors. Inspired by this fact, a global saliency model is constructed based on high-frequency coefficients of the multi-scale and multi-direction wavelet decomposition, which can characterize different feature information from edge to texture of the input image. To further reduce the false alarms, a new and effective multi-level discrimination method is designed based on the improved entropy and pixel distribution, which is robust against the interferences introduced by islands, coastlines, clouds, and shadows. The experimental results on optical remote sensing images validate that the presented saliency model outperforms the comparative models in terms of the area under the receiver operating characteristic curves core and the accuracy in the images with different sizes. After the target identification, the locations and the number of the ships in various sizes and colors can be detected accurately and fast with high robustness.

  17. Alarm management for storage and transportation terminals; Gerenciamento de alarmes para terminais de transferencia e estocagem

    Energy Technology Data Exchange (ETDEWEB)

    Loureiro, Patricia [PETROBRAS, Rio de Janeiro, RJ (Brazil); Feldman, Rafael Noac [PETROBRAS Transporte S.A. (TRANSPETRO), Rio de Janeiro, RJ (Brazil)

    2005-07-01

    Recently, in many industrial segments, it has been taken into account the issues related to the high amount of alarms that are announced in the control rooms, even if the industrial process is under normal conditions. Recent studies and surveys have shown that the three major problems related to it are: alarms that remain active during normal operation; alarms that remain chattering during an operational period; the phenomenon called Alarm flood, that occurs when an extensive amount of alarms is announced and the operator does not have enough time to take effective actions. In order to reduce or to eliminate the two above mentioned causes, alarm analysis and housekeeping, called Alarm Rationalization, have been efficient in major cases, because such facts occur mainly due to inadequate limits definition and/or equipment and instruments out of service or in maintenance. Such alarms are called in the literature as bad-actors or villains, and their occurrences may reach up to 50% of the daily total amount of alarms. This paper aims to present the main results of a project named Alarm Management for Transfer and Storage Terminals. The project development is based on two different terminal surveys, in order not only to identify the most frequent causes of undesirable alarms, but also to generate design standards. The main phases of the project are: alarm rationalization based on bad-actors detection; generate a set of design and operation standards; generate an Alarm Philosophy document for the Terminals. (author)

  18. Bilevel alarm monitoring multiplexer

    International Nuclear Information System (INIS)

    Johnson, C.S.

    1977-06-01

    This report describes the operation of the Bilevel Alarm Monitoring Multiplexer used in the Adaptive Intrusion Data System (AIDS) to transfer and control alarm signals being sent to the Nova 2 computer, the Memory Controlled Data Processor, and its own integral Display Panel. The multiplexer can handle 48 alarm channels and format the alarms into binary formats compatible with the destination of the alarm data

  19. Alarm system for ABWR main control panels

    Energy Technology Data Exchange (ETDEWEB)

    Kobayashi, Yuji; Saito, Koji [Toshiba Corp., Yokohoma (Japan)

    1997-09-01

    TOSHIBA has developed integrated digital control and instrumentation system for ABWR, which is the third-generation man machine interface system for main control room that we call A-PODIA (Advanced PODIA). A-Podia has been introduced the first actual ABWR plant in Japan. in A-PODIA, TOSHIBA has realized improvement of alarm system that all operator crews in the control room can recognize plant anomalies easily. The alarm system can recognize essential alarms for plant safety easily and understand annunciators with each integrated annunciators and their prioritized color easily by classifying alarms into plant-level essential annunciators, system-level integrated annunciators and equipment level individual annunciators with hierarchical structure. This paper describes conventional alarm system and the design philosophy, alarm system design and operation of ``Alarm System for ABWR Main Control Panels``. (author). 5 refs, 8 figs, 1 tab.

  20. Alarm system for ABWR main control panels

    International Nuclear Information System (INIS)

    Kobayashi, Yuji; Saito, Koji

    1997-01-01

    TOSHIBA has developed integrated digital control and instrumentation system for ABWR, which is the third-generation man machine interface system for main control room that we call A-PODIA (Advanced PODIA). A-Podia has been introduced the first actual ABWR plant in Japan. in A-PODIA, TOSHIBA has realized improvement of alarm system that all operator crews in the control room can recognize plant anomalies easily. The alarm system can recognize essential alarms for plant safety easily and understand annunciators with each integrated annunciators and their prioritized color easily by classifying alarms into plant-level essential annunciators, system-level integrated annunciators and equipment level individual annunciators with hierarchical structure. This paper describes conventional alarm system and the design philosophy, alarm system design and operation of ''Alarm System for ABWR Main Control Panels''. (author). 5 refs, 8 figs, 1 tab

  1. REACH: an effective catalyst for scaling up priority nutrition interventions at the country level.

    Science.gov (United States)

    Pearson, Brenda L; Ljungqvist, Björn

    2011-06-01

    Renewed Efforts Against Child Hunger (REACH) is the joint United Nations initiative to address Millennium Development Goal (MDG) 10, Target 3, i.e., to halve the proportion of underweight children under 5 years old by 2015. The United Nations Food and Agriculture Organization (FAO), the World Health Organization (WHO), the United Nations Children's Fund (UNICEF), the World Food Programme (WFP), and the International Fund for Agricultural Development (IFAD) developed and tested a facilitation mechanism to act as a catalyst for scaling up multisectoral nutrition activities. The UN-REACH partners developed pilot projects in Mauritania and Lao PDR from 2008 to 2010 and deployed facilitators to improve nutrition governance and coordination. Review missions were conducted in February 2011 to assess the REACH approach and what it achieved. The UN review mission members reviewed documents, assessed policy and management indicators, conducted qualitative interviews, and discussed findings with key stakeholders, including the most senior UN nutrition directors from all agencies. Among other UN-REACH achievements, the Prime Minister of Mauritania agreed to preside over a new National Nutrition Development Council responsible for high-level decision-making and setting national policy objectives. REACH facilitated the completion of Lao's first national Nutrition Strategy and Plan of Action and formation of the multistakeholder Nutrition Task Force. During the REACH engagement, coordination, joint advocacy, situation analysis, policy development, and joint UN programming for nutrition were strengthened in Lao PDR and Mauritania. Improvements in the nutrition governance and management mechanisms in Mauritania and Lao PDR were observed during the period of REACH support through increased awareness of nutrition as a key development objective, establishment of governmental multisectoral coordinating mechanisms, improved government capacity, and new joint UN-government nutrition

  2. Ionization smoke detector and alarm system

    International Nuclear Information System (INIS)

    1974-01-01

    An ionization smoke detector particularly suited to residential use is disclosed. The detector is battery-operated and is connected with a non-latching, pulsating alarm circuit. The detector has a sensing chamber formed by a perforated metallic shell and an electrode within which an insulated radiation source is centrally positioned to generate an ionization current for detecting smoke or other similar aerosols. The alarm circuit provides a pulsating alarm signal when smoke levels above a pre-determined value are sensed. The alarm circuit also includes a low voltage detection circuit for sounding the alarm when the end of useful battery life is approaching. (Auth.)

  3. Alarm pocket dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Hiraki, H; Kitamura, S [Matsushita Electric Industrial Co. Ltd., Kadoma, Osaka (Japan)

    1975-04-01

    This instrument is a highly reliable pocket dosimeter which has been developed for personal monitoring use. The dosimeter generates an alarm sound when the exposure dose reaches a preset value. Using a tiny GM tube for a radiation detector and measuring the integrated dose by means of a digital counting method, this new pocket dosimeter has high accuracy and stability. Using a sealed alkali storage battery for the power supply, and with an automatic control charger, this dosimetry system is easy and economical to operate and maintain. Detectable radiation by the dosimeter are X and ..gamma.. rays. Standard preset dose values are 30, 50, 80 and 100 mR. Detection accuracy is betwen +10% and -20%. The dosimeter is continuously usable for more than 14 hours after charging for 2 hours. The dosimeter has the following features; good realiability, shock-proof loud and clear alarm sound, the battery charger also serves as a stock container for the dosimeters, and no switching operation required for the power supply due to the internal automatic switch. Therefore, the dosimetry system is very useful for personal monitoring management in many radiation industry establishments.

  4. For your eyes only: Effect of confederate's eye level on reach-to-grasp action

    Directory of Open Access Journals (Sweden)

    Francois eQuesque

    2014-12-01

    Full Text Available Previous studies have shown that the spatio-temporal parameters of reach-to-grasp movement are influenced by the social context in which the motor action is performed. In particular, when interacting with a confederate, movements are slower, with longer initiation times and more ample trajectories, which has been interpreted as implicit communicative information emerging through voluntary movement to catch the partner’s attention and optimize cooperation (Quesque et al., 2013. Because gaze is a crucial component of social interactions, the present study evaluated the role of a confederate's eye level on the social modulation of trajectory curvature. An actor and a partner facing each other took part in a cooperative task consisting, for one of them, of grasping and moving a wooden dowel under time constraints. Before this Main action, the actor performed a Preparatory action, which consisted of placing the wooden dowel on a central marking. The partner's eye level was unnoticeably varied using an adjustable seat that matched or was higher than the actor’s seat. Our data confirmed the previous effects of social intention on motor responses. Furthermore, we observed an effect of the partner's eye level on the Preparatory action, leading the actors to exaggerate unconsciously the trajectory curvature in relation to their partner's eye level. No interaction was found between the actor's social intention and their partner's eye level. These results suggest that other bodies are implicitly taken into account when a reach-to-grasp movement is produced in a social context.

  5. Implementation of the Integrated Alarm System for KOMAC facility using EPICS framework and Eclipse

    International Nuclear Information System (INIS)

    Song, Young-Gi; Kim, Jae-Ha; Kim, Han-Sung; Kwon, Hyeok-Jung; Cho, Yong-Sub

    2017-01-01

    The alarm detecting layer is the component that monitors alarm signals which are transported to the processing part through message queue. The main purpose of the processing part is to transfer the alarm signals connecting an alarm identification and state of the alarm to database system. The operation interface of system level signal links has been developed by EPICS framework. EPICS tools have been used for monitoring device alarm status. The KOMAC alarm system was developed for offering a user-friendly, intuitive user interface. The alarm system is implemented with EPICS IOC for alarm server, eclipse-mars integrated development tool for alarm viewer, and mariadb for alarm log. The new alarm system supports intuitive user interface on alarm information and alarm history. Alarm view has plans to add login function, user permission on alarm acknowledge, user permission of PV import, search and report function.

  6. Improving the safety of Ukrainian NPP to reach an internationally accepted level

    International Nuclear Information System (INIS)

    Bozhko, S.; Helske, J.; Janke, R.; Mayoral, C.

    2013-01-01

    This paper summarizes the safety status and the modernization progress of Ukrainian NPPs towards an internationally accepted level of safety. After a brief discussion of the concept of what is called an 'international accepted level' for new and operating NPPs, the status of Russian type WWER and in particular the Ukrainian NPPs is presented. Then, the performed investigations of the gaps between international accepted level and the original status of Ukrainian NPPs are presented. The safety objectives of the modernization programs, some examples of defence in depth improvements, and an overall view of the modernization programs of Ukrainian NPPs are produced. Then, few important safety improvements implemented at the oldest Ukrainian WWER-1000 South Ukraine-1 are given in more detail. Finally, a conclusion presents the current status on the way to fulfill the national safety targets and to reach an internationally accepted level for all the Ukrainian NPPs. The paper is followed by the slides of the presentation. (authors)

  7. Time to reach a given level of number of neutrons is stochastic analog of reactor period

    International Nuclear Information System (INIS)

    Ryazanov, V.V.

    2012-01-01

    In theory and in practice the operation of nuclear reactors to control the safety of the reactor is widely used deterministic value - the period of the reactor. It is proposed along with the period of the reactor using a stochastic analogue of this magnitude - a random amount of time to achieve a given level of a random process for the number of neutrons in the reactor. The paper discusses various features of the behavior of the mean and variance of time to achieve a specified level. This kind of features can be associated with impaired behavior of the reactor system. Introduced the value of time required to reach the level can be used to monitor and improve the safety of nuclear power plants

  8. Alarm-Processing in Nuclear Power Plants

    International Nuclear Information System (INIS)

    Otazo, J; Fernandez, R

    2000-01-01

    Information overload due to the activation of a great number of alarms in a short time is a common problem for the operator in the control room of a industrial plant, mainly in complex process like the nuclear power plants.The problem is the conventional conception of the alarm system, that defines each alarm like a separated and independent entity of the global situation of the plant.A direct consequence is the generation of multiple alarms during a significative disturbance in the process, being most of them redundant and irrelevant to the actual process state wich involves an extra load to the operator, who wastes time in acting selecting the important alarms of the group that appears or lead to a an erroneous action.The present work first describes the techniques developed in the last years to attack the avalanche of alarms problem.Later we present our approach to alarm-processing: an expert system as alarm-filter.Our objective is collect in the system the state of the art in the development of advanced alarm systems, offering an improvement of the information flow to the operators through the suppression of nonsignificant alarms and a structured visualization of the process state.Such support is important during a disturbance for the identification of plant state, diagnosis, consequence prediction and corrective actions.The system is arranged in three stages: alarm-generation, alarm-filter and alarm-presentation.The alarm-generation uses conventional techniques or receives them from an external system.The alarm-filter uses suppression techniques based on: irrelevance analysis with the operation mode and the state of components, causal reasoning and static importance analysis.The alarm presentation is made through a structured way using a priority scheme with three level.The knowledge representation of each alarm is based on frames and a graph of alarms for global knowledge, where the connections between nodes represent causal and irrelevance relations

  9. Assessment of impacts from water level fluctuations on fish in the Hanford Reach, Columbia River

    Energy Technology Data Exchange (ETDEWEB)

    Becker, C.D.; Fickeisen, D.H.; Montgomery, J.C.

    1981-05-01

    Observations on the effects of water level fluctuations in the Hanford Reach of the Columbia River, Washington, were made in 1976 and 1977. The two years provided contrasting flow regimes: high water and fluctuations of greater magnitude prevailed in 1976; low water and higher temperatures prevailed in 1977. Situations where fish and other aquatic organisms were destroyed by changing water levels were observed and evaluated each year in three study areas: Hanford, F-Area, and White Bluffs sloughs. Losses primarily were due to stranding, entrapment (with or without complete dewatering), and predation. Juvenile fish were more susceptible to entrapment and stranding than were adult fish. Estimates of actual losses were biased and conservative because relatively few fish could be found after each decline of water level and dewatering. The most valued species of fish affected by water level fluctuations at Hanford were the anadromus fall chinook salmon (Oncorhynchus tshawytscha) and the resident smallmouth bass (Micropterus dolomieui). Crucial periods for chinook salmon occurred during winter when incubating eggs were in the gravel of the main channel, and before and during seaward migration in the spring when fry were abundant in shoreline zones. The crucial period for smallmouth bass was during spring and early summer when adults were spawning in warmed sloughs and shoreline zones. Chinook salmon and smallmouth bass fry were vulnerable to stranding and entrapment, and smallmouth bass nests were susceptible to exposure and temperature changes resulting from repeated water level fluctuations. Thus, flow manipulation may be crucial to their survival. The extent to which other species of riverine fish were affected by water level fluctuations depended upon their use of shoreline zones for spawning and rearing young.

  10. A distributed water level network in ephemeral river reaches to identify hydrological processes within anthropogenic catchments

    Science.gov (United States)

    Sarrazin, B.; Braud, I.; Lagouy, M.; Bailly, J. S.; Puech, C.; Ayroles, H.

    2009-04-01

    In order to study the impact of land use change on the water cycle, distributed hydrological models are more and more used, because they have the ability to take into account the land surface heterogeneity and its evolution due to anthropogenic pressure. These models provide continuous distributed simulations of streamflow, runoff, soil moisture, etc, which, ideally, should be evaluated against continuous distributed measurements, taken at various scales and located in nested sub-catchments. Distributed network of streamflow gauging stations are in general scarce and very expensive to maintain. Furthermore, they can hardly be installed in the upstream parts of the catchments where river beds are not well defined. In this paper, we present an alternative to these standard streamflow gauging stations network, based on self powered high resolution water level sensors using a capacitive water height data logger. One of their advantages is that they can be installed even in ephemeral reaches and from channel head locations to high order streams. Furthermore, these innovative and easily adaptable low cost sensors offer the possibility to develop in the near future, a wireless network application. Such a network, including 15 sensors has been set up on nested watersheds in small and intermittent streams of a 7 km² catchment, located in the mountainous "Mont du Lyonnais" area, close to the city of Lyon, France. The land use of this catchment is mostly pasture, crop and forest, but the catchment is significantly affected by human activities, through the existence of a dense roads and paths network and urbanized areas. The equipment provides water levels survey during precipitation events in the hydrological network with a very accurate time step (2 min). Water levels can be related to runoff production and catchment response as a function of scale. This response will depend, amongst other, on variable soil water storage capacity, physiographic data and characteristics of

  11. Display-And-Alarm Circuit For Accelerometer

    Science.gov (United States)

    Bozeman, Richard J., Jr.

    1995-01-01

    Compact accelerometer assembly consists of commercial accelerometer retrofit with display-and-alarm circuit. Provides simple means for technician attending machine to monitor vibrations. Also simpifies automatic safety shutdown by providing local alarm or shutdown signal when vibration exceeds preset level.

  12. Determination of nitrogen reduction levels necessary to reach groundwater quality targets in Slovenia.

    Science.gov (United States)

    Andelov, Miso; Kunkel, Ralf; Uhan, Jože; Wendland, Frank

    2014-09-01

    Within a collaborative project between Slovenian Environment Agency (ARSO) and Research Center Jülich (FZJ), nitrogen reduction levels necessary to reach groundwater quality targets in Slovenia were assessed. For this purpose the hydrological model GROWA-DENUZ was coupled with agricultural N balances and applied consistently to the whole territory of Slovenia in a spatial resolution of 100×100m. GROWA was used to determine the water balance in Slovenia for the hydrologic period 1971-2000. Simultaneously, the displaceable N load in soil was assessed from agricultural Slovenian N surpluses for 2011 and the atmospheric N deposition. Subsequently, the DENUZ model was used to assess the nitrate degradation in soil and, in combination with the percolation water rates from the GROWA model, to determine nitrate concentration in the leachate. The areas showing predicted nitrate concentrations in the leachate above the EU groundwater quality standard of 50mg NO3(-)/L have been identified as priority areas for implementing nitrogen reduction measures. For these "hot spot" areas DENUZ was used in a backward mode to quantify the maximal permissible nitrogen surplus levels in agriculture to guarantee a nitrate concentration in percolation water below 50mg NO3(-)/L. Model results indicate that additional N reduction measures should be implemented in priority areas rather than area-covering. Research work will directly support the implementation of the European Union Water Framework Directive in Slovenia, e.g., by using the maximal permissible nitrogen surplus levels as a framework for the derivation of regionally adapted and hence effective nitrogen reduction measures. Copyright © 2014. Published by Elsevier B.V.

  13. Transportable criticality alarm system

    International Nuclear Information System (INIS)

    Clem, W.E.

    1988-09-01

    The Transportable Criticality Alarm System was developed at the Hanford Site in 1982 to comply with the requirements of US Department of Energy Order DOE 5480.1, 12/18/80, and ANSI/ANS-8.3- 1979. The portable unit that it replaced failed to comply with the new requirements in that it did not provide the necessary warning of malfunctions, nor did it provide the Hanford Site standard criticality alarm signal. Modern technology allowed the Transportable Criticality Alarm System to comply with the criticality requirements cited and to incorporate other features that make it more usable, maintainable, and reliable. The Transportable Criticality Alarm System (TCAS) provides temporary criticality coverage in manned areas where the facility criticality alarm system is not operable. This gamma radiation-sensitive system has been in use for the past 6 yr at the Hanford Site. 2 refs., 4 figs., 1 tab

  14. Functional alarming and information retrieval

    International Nuclear Information System (INIS)

    Goodstein, L.P.

    1985-08-01

    This paper deals with two facets of the design and efficient utilisation by operating personnel of computer-based interfaces for monitoring and the supervisory control of complex industrial systems - e.g., power stations, chemical plants, etc. These are alarming and information retrieval both of which are extremely sensitive to computerisation. For example, the advent of computers for display requires that some means of assuring easy and rapid access to large amounts of relevant stored information be found. In this paper, alarming and information retrieval are linked together through a multilevel functional description of the target plant. This representation serves as a framework for structuring the access to information as well as defining associated ''alarms'' at the various descriptive levels. Particular attention is paid to the level where mass and energy flows and balances are relevant. It is shown that the number of alarms here is reduced considerably while information about content and interrelationships is enhanced - which at the same time eases the retrieval problem. (author)

  15. Alarm handling systems and techniques developed to match operator tasks

    Energy Technology Data Exchange (ETDEWEB)

    Bye, A; Moum, B R [Institutt for Energiteknikk, Halden (Norway). OECD Halden Reaktor Projekt

    1997-09-01

    This paper covers alarm handling methods and techniques explored at the Halden Project, and describes current status on the research activities on alarm systems. Alarm systems are often designed by application of a bottom-up strategy, generating alarms at component level. If no structuring of the alarms is applied, this may result in alarm avalanches in major plant disturbances, causing cognitive overload of the operator. An alarm structuring module should be designed using a top-down approach, analysing operator`s tasks, plant states, events and disturbances. One of the operator`s main tasks during plant disturbances is status identification, including determination of plant status and detection of plant anomalies. The main support of this is provided through the alarm systems, the process formats, the trends and possible diagnosis systems. The alarm system should both physically and conceptually be integrated with all these systems. 9 refs, 5 figs.

  16. Alarm handling systems and techniques developed to match operator tasks

    International Nuclear Information System (INIS)

    Bye, A.; Moum, B.R.

    1997-01-01

    This paper covers alarm handling methods and techniques explored at the Halden Project, and describes current status on the research activities on alarm systems. Alarm systems are often designed by application of a bottom-up strategy, generating alarms at component level. If no structuring of the alarms is applied, this may result in alarm avalanches in major plant disturbances, causing cognitive overload of the operator. An alarm structuring module should be designed using a top-down approach, analysing operator's tasks, plant states, events and disturbances. One of the operator's main tasks during plant disturbances is status identification, including determination of plant status and detection of plant anomalies. The main support of this is provided through the alarm systems, the process formats, the trends and possible diagnosis systems. The alarm system should both physically and conceptually be integrated with all these systems. 9 refs, 5 figs

  17. Alarm criteria for the fixed gamma radiation monitoring stations

    International Nuclear Information System (INIS)

    Kjelle, P.E.

    1987-01-01

    For the type of measurement considered here, an alarm based on the dose excess has been shown to be a more certain method of indicating that a true alarm situation exists than an alarm based on the dose-rate exceeding a given dose-rate level. The method offers a number of advantages: Dose-rate levels can be recorded over long periods without any necessity of making manual changes in the alarm level due to seasonal variations in the background level. A relatively low level can be specified without an unnecessary number of false alarms being triggered by gamma-emitting radon daughters. (orig./HP)

  18. Advanced alarm management system

    International Nuclear Information System (INIS)

    Easter, J.R.

    1995-01-01

    The Westinghouse Advanced Alarm Management System (AWARE) is one of the Man-Machine Design Interfaces (MMI) which has great flexibility with regard to hardware type and configuration, alarm system concept, plant scope, engineering scope and installation. The AWARE System provides the capability to better manage the quantity prioritization and presentation of real-time process alarm messages in the control room. The messages are specific, precise and dynamic. The AWARE System can provide a large reduction in the number of messages that the control room staff must address at any one time, thus making the alarm message system a useful tool for the operators during situations that normally produce a high volume of messages as well as improving the clarity of the presentation of process abnormalities during small disturbances. The operating staff is now provided with the basis for a better understanding of the current plant state and for taking the appropriate control actions. (2 refs., 3 figs.)

  19. Speech Alarms Pilot Study

    Science.gov (United States)

    Sandor, Aniko; Moses, Haifa

    2016-01-01

    Speech alarms have been used extensively in aviation and included in International Building Codes (IBC) and National Fire Protection Association's (NFPA) Life Safety Code. However, they have not been implemented on space vehicles. Previous studies conducted at NASA JSC showed that speech alarms lead to faster identification and higher accuracy. This research evaluated updated speech and tone alerts in a laboratory environment and in the Human Exploration Research Analog (HERA) in a realistic setup.

  20. Security Alarm for Motorcycle

    Directory of Open Access Journals (Sweden)

    Ismail Ismail

    2015-11-01

    Full Text Available A security alarm for motorcycle has been developed. This equipment consists of two parts. Part one is as a remote control, where it produces a radio signal with frequency of 37.5 MHz to turn on (activate or to turn off alarm. Part two consists of sensor, receiver to receive signal from part one, and alarm. This part two will be attached to motorcycle while part one will be kept as a key by the owner of motorcycle. This equipment has been tested in the laboratory and it worked well. When part two is activated by pushing the “set button” in part one, then any movement of part two (as a movement of motor cycle by about 20 cm from initial position will cause the alarm sounds continuously. The alarm will be off whenever the “reset button” in part one is pushed. Part one (a remote control can activate part two with a maximum of twelve meter separation apart. This shows that the equipment can be used as a security alarm to prevent the motorcycle to be stolen in the future.

  1. Interior intrusion alarm systems

    International Nuclear Information System (INIS)

    Prell, J.A.

    1978-01-01

    In meeting the requirements for the safeguarding of special nuclear material and the physical protection of licensed facilities, the licensee is required to design a physical security system that will meet minimum performance requirements. An integral part of any physical security system is the interior intrusion alarm system. The purpose of this report is to provide the potential user of an interior intrusion alarm system with information on the various types, components, and performance capabilities available so that he can design and install the optimum alarm system for his particular environment. In addition, maintenance and testing procedures are discussed and recommended which, if followed, will help the user obtain the optimum results from his system

  2. The LEP alarm system

    International Nuclear Information System (INIS)

    Tyrrell, M.W.

    1992-01-01

    Unlike alarm systems for previous accelerators, the LEP alarm system caters not only for the operation of the accelerator but also for technical services and provides the direct channel for personnel safety. It was commissioned during 1989 and has seen a continued development up to the present day. The system, comprising over 50 computers including 5 different platforms and 4 different operating systems, is described. The hierarchical structure of the software is outlined from the interface to the equipment groups, through the front end computers to the central server, and finally to the operator consoles. Reasons are given for choosing a conventional, as opposed to a 'knowledge based' approach. Finally, references are made to a prototype real time expert system for surveying the power converters of LEP, which was conducted during 1990 as part of the alarm development program. (author)

  3. Criticality accident alarm system

    International Nuclear Information System (INIS)

    Malenfant, R.E.

    1991-01-01

    The American National Standard ANSI/ANS-8.3-1986, Criticality Accident Alarm System provides guidance for the establishment and maintenance of an alarm system to initiate personnel evacuation in the event of inadvertent criticality. In addition to identifying the physical features of the components of the system, the characteristics of accidents of concern are carefully delineated. Unfortunately, this ANSI Standard has led to considerable confusion in interpretation, and there is evidence that the ''minimum accident of concern'' may not be appropriate. Furthermore, although intended as a guide, the provisions of the standard are being rigorously applied, sometimes with interpretations that are not consistent. Although the standard is clear in the use of absorbed dose in free air of 20 rad, at least one installation has interpreted the requirement to apply to dose in soft tissue. The standard is also clear in specifying the response to both neutrons and gamma rays. An assembly of uranyl fluoride enriched to 5% 235 U was operated to simulate a potential accident. The dose, delivered in a free run excursion 2 m from the surface of the vessel, was greater than 500 rad, without ever exceeding a rate of 20 rad/min, which is the set point for activating an alarm that meets the standard. The presence of an alarm system would not have prevented any of the five major accidents in chemical operations nor is it absolutely certain that the alarms were solely responsible for reducing personnel exposures following the accident. Nevertheless, criticality alarm systems are now the subject of great effort and expense. 13 refs

  4. Utilizing Facebook and Automated Telephone Calls to Increase Adoption of a Local Smoke Alarm Installation Program.

    Science.gov (United States)

    Frattaroli, Shannon; Schulman, Eric; McDonald, Eileen M; Omaki, Elise C; Shields, Wendy C; Jones, Vanya; Brewer, William

    2018-05-17

    Innovative strategies are needed to improve the prevalence of working smoke alarms in homes. To our knowledge, this is the first study to report on the effectiveness of Facebook advertising and automated telephone calls as population-level strategies to encourage an injury prevention behavior. We examine the effectiveness of Facebook advertising and automated telephone calls as strategies to enroll individuals in Baltimore City's Fire Department's free smoke alarm installation program. We directed our advertising efforts toward Facebook users eligible for the Baltimore City Fire Department's free smoke alarm installation program and all homes with a residential phone line included in Baltimore City's automated call system. The Facebook campaign targeted Baltimore City residents 18 years of age and older. In total, an estimated 300 000 Facebook users met the eligibility criteria. Facebook advertisements were delivered to users' desktop and mobile device newsfeeds. A prerecorded message was sent to all residential landlines listed in the city's automated call system. By the end of the campaign, the 3 advertisements generated 456 666 impressions reaching 130 264 Facebook users. Of the users reached, 4367 individuals (1.3%) clicked the advertisement. The automated call system included approximately 90 000 residential phone numbers. Participants attributed 25 smoke alarm installation requests to Facebook and 458 to the automated call. Facebook advertisements are a novel approach to promoting smoke alarms and appear to be effective in exposing individuals to injury prevention messages. However, converting Facebook message recipients to users of a smoke alarm installation program occurred infrequently in this study. Residents who participated in the smoke alarm installation program were more likely to cite the automated call as the impetus for their participation. Additional research is needed to understand the circumstances and strategies to effectively use the social

  5. Blood corticosterone concentration reaches critical illness levels early during acute malnutrition in the weanling mouse.

    Science.gov (United States)

    Monk, Jennifer M; Makinen, Kimberley; Shrum, Bradly; Woodward, Bill

    2006-03-01

    Acute (i.e., wasting) pediatric malnutrition consistently elevates blood glucocorticoid levels, but neither the magnitude of the rise in concentration nor its kinetics is clear. Male and female C57BL/6J mice, initially 19 days old, and CBA/J mice, initially 23 days old, consumed a complete purified diet either ad libitum (age-matched control) or in restricted daily quantities (mimicking marasmus), or they consumed a purified isocaloric low-protein diet ad libitum (mimicking incipient kwashiorkor). Serum levels of corticosterone were assessed by double antibody radioimmunoassay after 3, 6, and 14 days (C57BL/6J strain) or after 6 and 14 days in the genetically distant CBA/J strain. Age-matched control groups of both strains exhibited mean corticosterone levels of 5-30 ng/ml, whereas the acutely malnourished groups exhibited mean levels of this hormone that were elevated by more than an order of magnitude as early as 3 days after initiation of weight loss. This outcome was confirmed in a second experiment in which the serum corticosterone level of C57BL/6J weanlings was examined by competitive binding enzyme immunoassay 3 and 14 days after initiation of the dietary protocols. Therefore, deficits of protein and/or energy in weanling murine systems relevant to acute pediatric malnutrition elicit early elevations in blood glucocorticoid levels to a magnitude reminiscent of critical illness and multiple trauma. The key to this novel finding was an exsanguination method that permitted accurate assessment of the blood corticosterone level of the healthy, quiescent mouse. Overall, the results of this investigation provide a new perspective on the glucocorticoids as part of the early hormonal response to acute weanling malnutrition coincident with the shift toward catabolic metabolism and the initiation of depression in cellular immune competence.

  6. Why Do At-Risk Mothers Fail To Reach Referral Level? Barriers ...

    African Journals Online (AJOL)

    In southern Tanzania, few high-risk pregnancies are channeled through antenatal care to the referral level. We studied the influences that make pregnant women heed or reject referral advice. Semi-structured interviews with sixty mothers-to-be, twenty-six health workers and six key-informants to identify barriers to use of ...

  7. Dynamic alarm response procedures

    International Nuclear Information System (INIS)

    Martin, J.; Gordon, P.; Fitch, K.

    2006-01-01

    The Dynamic Alarm Response Procedure (DARP) system provides a robust, Web-based alternative to existing hard-copy alarm response procedures. This paperless system improves performance by eliminating time wasted looking up paper procedures by number, looking up plant process values and equipment and component status at graphical display or panels, and maintenance of the procedures. Because it is a Web-based system, it is platform independent. DARP's can be served from any Web server that supports CGI scripting, such as Apache R , IIS R , TclHTTPD, and others. DARP pages can be viewed in any Web browser that supports Javascript and Scalable Vector Graphics (SVG), such as Netscape R , Microsoft Internet Explorer R , Mozilla Firefox R , Opera R , and others. (authors)

  8. Toxicity alarm: Case history

    International Nuclear Information System (INIS)

    Hogan, D.; Retallack, J.

    1993-01-01

    In late fall 1991, the Novacor petrochemical plant near Joffre, Alberta experienced a toxicity alarm, the first since its startup 14 years ago. Fish exposed to a normal toxicity test were stressed within 2 h and showed 100% mortality after 24 h. A history of the events leading up to, during, and after the toxicity alarm is presented. The major effluent sources were three cooling water systems. Although these sources are well characterized, the event causes were not immediately clear. Initial toxic screening indicated that one was very toxic, another moderately toxic, and the third not toxic at all. All three systems utilized the same chemical treatment program to avoid fouling: stabilized phosphates with minor variants. The most toxic of the cooling systems operated at 10-12 cycles, had three chemicals for biocide control, and had three makeup streams. Toxic and nontoxic system characteristics were compared. An in-depth modified toxicity identification and evaluation program was then performed to identify and evaluate the cause of the toxicity alarm for future prevention. The most probable causes of toxicity were identified by elimination. The combination of high numbers of cycles, hydrocarbons in the makeup water, and bromine added as an antifoulant resulted in formation of aromatic bromamines which are capable of causing the toxic condition experienced. 2 tabs

  9. Ambition to reach zero level failure in VVER 1000 with russian fuel

    International Nuclear Information System (INIS)

    Mečíř, V.

    2015-01-01

    The purpose of “The Zero Failure Level Project” is to bring to real operation of VVER 1000 units the dream of all utilities such as problem free and cost effective operation. This essentially turns into requirement on failure free fuel operation. At the same time the general requirements such as safety, cost effectiveness, operational flexibility, fuel cycle and fuel flexibility need to be satisfied. Several specific tasks were performed and many of them are still in process. Specific failure tree was developed in a format, which allows step by step failure tree improvement. Fuel types and its modifications, taking into account manufacturing conditions, were specified. In parallel with fuel types classification, real operational conditions were evaluated based on approximately 280 parameters by fuel assembly design features, operational procedures and practices and about 250 reactor unit parameters. As a result of this stage, groups of units with similar fuel operational conditions should be revealed and experience sharing database created. It is also recognized a need for consistent methods of operational data and data from pool side fuel assembly inspection. In the area of Foreign Material Exclusion activities closer cooperation between utility and supplier should be established including foreign material classification and improvement in root cause investigation

  10. Alarm handler for the advanced photon source control system

    International Nuclear Information System (INIS)

    Kraimer, M.R.; Cha, B.K.; Anderson, M.

    1991-01-01

    The Advanced Photon Source (APS), now under construction at Argonne National Laboratory, will have a control system employing graphics workstations at the operator interface level and VME-based microprocessors operating with a distributed database at the field level. The alarm handler is an application utilizing X-Windows running on one or more operator interface workstations which monitors alarms generated by the VME-based microprocessors. Alarms can be grouped in a hierarchical manner. The operator can monitor, acknowledge, and mask alarms either individually or aggregately. Alarm changes of state and all operator modifications are logged. When alarms occur, display windows are automatically generated conveying system and subsystem relationships and severity. Menus are used to modify the alarm action configuration files and to obtain help. Since alarm groups are defined via an alarm configuration file, the alarm handler is a general purpose application which can be customized to monitor a single subsystem or configured to monitor the entire accelerator complex. 2 refs., 2 figs

  11. Alarm personal dosemeter

    International Nuclear Information System (INIS)

    Nunomiya, Tomoya; Yamauchi, Hideshi; Shibata, Tetsuo

    2007-01-01

    Fuji Electric advanced the development of the conventional alarm personal dosemeter (APD) by adopting JIS Z 4312 (2002), which reflects IEC61526 (1998), and by improving the radiation performance characteristics, electromagnetic performance characteristics and mechanical performance characteristics of the APD (acquiring a CE marking). This new dosemeter will target overseas market in the future. The APD is a suitable dosemeter for detecting and monitoring radiation exposure to personnel working in the controlled areas of nuclear facilities, such as nuclear power stations. The APD product line includes 'gamma ray', 'gamma ray + beta ray' and 'gamma ray + neutron' models. (author)

  12. Speech Alarms Pilot Study

    Science.gov (United States)

    Sandor, A.; Moses, H. R.

    2016-01-01

    Currently on the International Space Station (ISS) and other space vehicles Caution & Warning (C&W) alerts are represented with various auditory tones that correspond to the type of event. This system relies on the crew's ability to remember what each tone represents in a high stress, high workload environment when responding to the alert. Furthermore, crew receive a year or more in advance of the mission that makes remembering the semantic meaning of the alerts more difficult. The current system works for missions conducted close to Earth where ground operators can assist as needed. On long duration missions, however, they will need to work off-nominal events autonomously. There is evidence that speech alarms may be easier and faster to recognize, especially during an off-nominal event. The Information Presentation Directed Research Project (FY07-FY09) funded by the Human Research Program included several studies investigating C&W alerts. The studies evaluated tone alerts currently in use with NASA flight deck displays along with candidate speech alerts. A follow-on study used four types of speech alerts to investigate how quickly various types of auditory alerts with and without a speech component - either at the beginning or at the end of the tone - can be identified. Even though crew were familiar with the tone alert from training or direct mission experience, alerts starting with a speech component were identified faster than alerts starting with a tone. The current study replicated the results from the previous study in a more rigorous experimental design to determine if the candidate speech alarms are ready for transition to operations or if more research is needed. Four types of alarms (caution, warning, fire, and depressurization) were presented to participants in both tone and speech formats in laboratory settings and later in the Human Exploration Research Analog (HERA). In the laboratory study, the alerts were presented by software and participants were

  13. Alarm management a comprehensive guide

    CERN Document Server

    Hollifield, Bill R

    2011-01-01

    In this second edition, Alarm Management: A Comprehensive Guide, various problems of alarm systems are covered with precise guidance on how they come about and how to effectively correct them. It is written by individuals with vast experience in the different plants, processes, and environments requiring effective alarm management. The second edition is filled with good examples and explanations of procedures, with practical lists and tips on how one should proceed. It is based on hundreds of successful projects.

  14. Fundamental Principles of Alarm Design

    DEFF Research Database (Denmark)

    Us, Tolga; Jensen, Niels; Lind, Morten

    2011-01-01

    Traditionally alarms are designed on the basis of empirical guidelines rather than on a sound scientific framework rooted in a theoretical foundation for process and control system design. This paper proposes scientific principles and a methodology for design of alarms based on a functional...... be applied to any engineering system which can be modeled by MFM. The methodology provides a set of alarms which can facilitate event interpretation and operator support for abnormal situation management. The proposed design methodology provides the information content of the alarms, but does not deal...

  15. Alarm management system

    International Nuclear Information System (INIS)

    Woods, D.D.; Elm, W.C.; Lipner, M.H.; Butterworth, G.E.; Easter, J.R.

    1989-01-01

    An alarm management system is described, comprising: a light water pressurized nuclear power plant; sensors coupled to the plant indicating the state of the plant; a sensor signal processor, operatively connected to the sensors, for producing state signals indicating the state of the sensors monitoring the power plant from the sensor signals and for combining the state signals using rule based algorithms to produce abnormality indication signals; a message processor, operatively connected to the sensor signal processor for receiving the abnormality indication signals; a spatially dedicated parallel display for each function, operatively connected to the message processor, for displaying the portion of the messages simultaneously; and a serial display, operatively connected to the message processor, for displaying the message in the queues in priority order within category within function upon request, the message processor outputting messages from the queue to the parallel display as display space becomes available on the parallel display due to an abnormality being resolved

  16. Habituating alarming atmospheres

    DEFF Research Database (Denmark)

    Højlund, Marie

    This paper proposes embodied rhythmic sound habituation as a possible resource when designing contextualized technologies in critical atmospheres. The main contribution is collating the concept of rhythm as presented by Henri Lefebvre with the concept of sound habituation to help operationalize...... functionality for the staff, but are stressful for visitors and patients, as they are designed to demand attention even though they have no direct functional meaning to them. By introducing sounds from the ward, integrated in the furniture as simple sound sample triggers, KidKit invites children to become...... accustomed to the alarming sounds through rhythmic interaction in the waiting room, and bringing the furniture with them afterwards as a secure anchor, when entering the ward. This rhythmic habituation can enable the child to focus her attention on the meeting with the hospitalized relative....

  17. Mental workload and motor performance dynamics during practice of reaching movements under various levels of task difficulty.

    Science.gov (United States)

    Shuggi, Isabelle M; Oh, Hyuk; Shewokis, Patricia A; Gentili, Rodolphe J

    2017-09-30

    The assessment of mental workload can inform attentional resource allocation during task performance that is essential for understanding the underlying principles of human cognitive-motor behavior. While many studies have focused on mental workload in relation to human performance, a modest body of work has examined it in a motor practice/learning context without considering individual variability. Thus, this work aimed to examine mental workload by employing the NASA TLX as well as the changes in motor performance resulting from the practice of a novel reaching task. Two groups of participants practiced a reaching task at a high and low nominal difficulty during which a group-level analysis assessed the mental workload, motor performance and motor improvement dynamics. A secondary cluster analysis was also conducted to identify specific individual patterns of cognitive-motor responses. Overall, both group- and cluster-level analyses revealed that: (i) all participants improved their performance throughout motor practice, and (ii) an increase in mental workload was associated with a reduction of the quality of motor performance along with a slower rate of motor improvement. The results are discussed in the context of the optimal challenge point framework and in particular it is proposed that under the experimental conditions employed here, functional task difficulty: (i) would possibly depend on an individuals' information processing capabilities, and (ii) could be indexed by the level of mental workload which, when excessively heightened can decrease the quality of performance and more generally result in delayed motor improvements. Copyright © 2017 IBRO. Published by Elsevier Ltd. All rights reserved.

  18. 46 CFR 28.250 - High water alarms.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 1 2010-10-01 2010-10-01 false High water alarms. 28.250 Section 28.250 Shipping COAST... Individuals On Board, or for Fish Tender Vessels Engaged in the Aleutian Trade § 28.250 High water alarms. On... operating station to indicate high water level in each of the following normally unmanned spaces: (a) A...

  19. Line supervision of alarm communications

    International Nuclear Information System (INIS)

    Chritton, M.R.

    1991-01-01

    The objective of this paper is to explain the role and application of alarm communication link supervision in security systems such as for nuclear facilities. The vulnerabilities of the various types of alarm communication links will be presented. Throughout the paper, an effort has been made to describe only those technologies commercially available and to avoid speculative theoretical solutions

  20. Alarm processing - Ways to the future

    Energy Technology Data Exchange (ETDEWEB)

    Pirus, D [EDF - Septen Service Etudes et Projets Thermiques et Nucleires, Villeurbanne Cedex (France)

    1997-09-01

    After a brief presentation of the main characteristics an efficient alarm system should have, a presentation of the N4 alarm processing and presentation is described in terms of reduction in alarm occurrence, alarm handling and operator presentation. The EDF experiments on the future alarm processing expected for the next generation of the French nuclear plants are then presented. This alarm system will manage the alarms functionally in order to present to the operators the real consequences on the whole plant of a dedicated alarm and try to imbed deeply the alarm presentation within the operating formats and the procedures. (author).

  1. Alarm processing - Ways to the future

    International Nuclear Information System (INIS)

    Pirus, D.

    1997-01-01

    After a brief presentation of the main characteristics an efficient alarm system should have, a presentation of the N4 alarm processing and presentation is described in terms of reduction in alarm occurrence, alarm handling and operator presentation. The EDF experiments on the future alarm processing expected for the next generation of the French nuclear plants are then presented. This alarm system will manage the alarms functionally in order to present to the operators the real consequences on the whole plant of a dedicated alarm and try to imbed deeply the alarm presentation within the operating formats and the procedures. (author)

  2. Development of radiation alarm monitor

    International Nuclear Information System (INIS)

    Myung Jae Song; Myung Chan Lee; Jung Kwan Son

    1997-01-01

    The Radiation Alarm Monitor is developed domestically in order to protect radiation workers from over exposure. The Radiation Alarm Monitor with microprocessor installed can record the information of radiation field before and after accidents. It can also provide the data to analyze the accident and to set a counterplan. It features a wide detection range of radiation (I OmR/h - I OOR/h), radiation work and data storage, portability, high precision (5%) due to calibration, and adaptation of a powerful alarm system. In order to protect workers from over exposure, light and sound alarm had been designed to initiate when accident occurs such as an unexpected change of radiation field such as radiation rate and accumulated dosed between 90 min. before the alarm and 30 min. after the alarm. In addition, the Radiation Alarm Monitor interfaces with computer so that the accident can be analyzed. After the testing conditions in other countries for the Radiation Alarm Monitor were compared, the most stringent test, ANSI N42. 17-A, was selected. The performance testing was car-ried out under various conditions of temperature, humidity, vibration and electromagnetic wave hindrance by Korea Research Institute of Standards and Science (KRISS). As a result, the Radiation Alan-n Monitor passed all test. Also, for the Radiation Alarm Monitor, environmental adaptability tests under the environmental conditions of NPP sites had been performed. The Radiation Alan-n Monitor had been reviewed by radiation workers at NPPs and their opinions had been collected. Operating procedure will be written and distributed to every NPP sites. Radiation Alarm Monitor will be modified for use under the specific environmental conditions of each site. It will be distributed to NPP sites and will be used by radiation workers

  3. Criticality alarm device

    International Nuclear Information System (INIS)

    Sasaki, Takashi.

    1995-01-01

    In a critical alarm device comprising an n-number of radiation detectors and a 2/3 logic module, the 2/3 logic module comprises a rectifier inputting an output of an AC power and outputting it as a full wave-rectified, DC voltage, an analog/digital converter inputting the output from the rectifier, putting it to analog/digital conversion and continuously storing it into a memory, a latch circuit inputting critical signals from the 2/3 logic and holding them and outputting latch signals, and a means for outputting timing signals for data storage to a memory based on clocks at a predetermined period and outputting timing signals for stopping the data storage to the memory after passing a predetermined period from the inputting time point if the latch signals are inputted from the latch circuit. Then, judgement whether the generation of criticality is caused by an actual radiation detection or by noises from the power source is enabled, to provide extremely high reliability. (N.H.)

  4. Transactions of the criticality alarm systems workshop

    International Nuclear Information System (INIS)

    1988-01-01

    The first Criticality Alarm workshop was held by the US Department of Energy Albuquerque Operations Office in 1985. This second workshop is the first held on an international level. There were 98 persons in attendance. They represented the Department of Energy (DOE) field offices, DOE contractors, the Nuclear Regulatory Commission (NRC), NRC licensees, and agencies in the United Kingdom, France, West Germany, and Japan. Topics were on practices experience, and development. A key value of the workshop was the sharing of critical alarm system experiences, problems, and advances in the state of the art. In addition, several Criticality Alarm Systems (CAS) equipment systems were exhibited. Papers were presented on: nature of criticality accidents; lessons learned from past accidents; application of ANS 8.3 standard; gamma and neutron detection systems; research and development in progress; testing at Oak Ridge and Los Alamos; methods used to place detectors; centralized readout feature; false alarms; trip-point settings; and testing and maintenance. The individual papers have been cataloged separately

  5. Poison and alarm: the Asian hornet Vespa velutina uses sting venom volatiles as an alarm pheromone.

    Science.gov (United States)

    Cheng, Ya-Nan; Wen, Ping; Dong, Shi-Hao; Tan, Ken; Nieh, James C

    2017-02-15

    In colonial organisms, alarm pheromones can provide a key fitness advantage by enhancing colony defence and warning of danger. Learning which species use alarm pheromone and the key compounds involved therefore enhances our understanding of how this important signal has evolved. However, our knowledge of alarm pheromones is more limited in the social wasps and hornets compared with the social bees and ants. Vespa velutina is an economically important and widespread hornet predator that attacks honey bees and humans. This species is native to Asia and has now invaded Europe. Despite growing interest in V. velutina , it was unknown whether it possessed an alarm pheromone. We show that these hornets use sting venom as an alarm pheromone. Sting venom volatiles were strongly attractive to hornet workers and triggered attacks. Two major venom fractions, consisting of monoketones and diketones, also elicited attack. We used gas chromatography coupled to electroantennographic detection (GC-EAD) to isolate 13 known and 3 unknown aliphatic ketones and alcohols in venom that elicited conspicuous hornet antennal activity. Two of the unknown compounds may be an undecen-2-one and an undecene-2,10-dinone. Three major compounds (heptan-2-one, nonan-2-one and undecan-2-one) triggered attacks, but only nonan-2-one did so at biologically relevant levels (10 hornet equivalents). Nonan-2-one thus deserves particular attention. However, the key alarm releasers for V. velutina remain to be identified. Such identification will help to illuminate the evolution and function of alarm compounds in hornets. © 2017. Published by The Company of Biologists Ltd.

  6. Some points of advanced alarm system design

    International Nuclear Information System (INIS)

    Hollo, E.

    1977-01-01

    A description of some of the more relevant questions relating to advanced alarm systems for nuclear power plant installations. The development of such alarm systems embodies three main tasks: development of formal alarm handling methods, design of alarm patterns, development of alarm analysis systems. The major aspects of these tests are dealt with and the close relation between the alarm analysis and the plant disturbance analysis procedure is emphasized. (author)

  7. [Effects of groundwater level on chlorophyll fluorescence characteristics of Tamarix hispida in lower reaches of Tarim River].

    Science.gov (United States)

    Zhu, Cheng-gang; Li, Wei-hong; Ma, Jian-xin; Ma, Xiao-dong

    2010-07-01

    Based on the monitoring data of groundwater level at the typical sections in lower reaches of Tarim River, three survey plots nearby the ecological monitoring wells with groundwater depths > 6 m were selected to investigate the chlorophyll fluorescence characteristics of Tamarix hispida and its photosynthetic activity of PSII under effects of different groundwater depths. With increasing groundwater depth, the chlorophyll fluorescence parameters such as actual photochemical efficiency of PSII in the light (phi(PSII)), electron transport rate (ETR), and photochemistry quenching (q(p)) of T. hispida decreased, while the non-photochemistry quenching (q(N), NPQ) and the yield for dissipation by down-regulation (Y(NPQ)) increased remarkably, and the maximal photochemical efficiency of PSII (Fv/Fm) maintained an optimum value. All the results suggested that the PSII photosynthetic activity of T. hispida under drought stress declined with increasing groundwater depth, and the greater excess energy could result in more risk of photo-inhibition. However, the good adaptability and drought tolerance of T. hispida could make its PSII not seriously damaged, though the drought stress actually existed.

  8. Ultrasonic Technology in Duress Alarms.

    Science.gov (United States)

    Lee, Martha A.

    2000-01-01

    Provides the pros and cons of the most commonly used technologies in personal duress alarm systems in the school environment. Discussed are radio frequency devices, infrared systems, and ultrasonic technology. (GR)

  9. Measurement of the Portsmouth Gaseous Diffusion Plant criticality accident alarm

    International Nuclear Information System (INIS)

    Tayloe, R.W. Jr.; McGinnis, B.

    1990-01-01

    Measurements of the Portsmouth Gaseous Diffusion Plant's nuclear criticality accident radiation alarm signal response time, sound wave frequency, and sound volume levels were made to demonstrate compliance with ANSI/ANS-8.3-1986. A steady-state alarm signal is produced within one-half second of obtaining a two-out-of-three detector trip. The fundamental alarm sound wave frequency is 440 hertz. The sound volume levels are greater than 10 decibels above background and ranged from 100 to 125 A-weighted decibels. The requirements of the standard were met; however the recommended maximum sound volume level of 115 dBA was exceeded. Emergency procedures require immediate evacuation upon initiation of a facility's radiation alarm. Comparison with standards for allowable time of exposure at different noise levels indicate that the elevated noise level at this location does not represent an occupational injury hazard. 8 refs., 5 figs

  10. Clinical Alarms in intensive care: implications of alarm fatigue for the safety of patients

    Directory of Open Access Journals (Sweden)

    Adriana Carla Bridi

    2014-12-01

    Full Text Available OBJECTIVES: to identify the number of electro-medical pieces of equipment in a coronary care unit, characterize their types, and analyze implications for the safety of patients from the perspective of alarm fatigue.METHOD: this quantitative, observational, descriptive, non-participatory study was conducted in a coronary care unit of a cardiology hospital with 170 beds.RESULTS: a total of 426 alarms were recorded in 40 hours of observation: 227 were triggered by multi-parametric monitors and 199 were triggered by other equipment (infusion pumps, dialysis pumps, mechanical ventilators, and intra-aortic balloons; that is an average of 10.6 alarms per hour.CONCLUSION: the results reinforce the importance of properly configuring physiological variables, the volume and parameters of alarms of multi-parametric monitors within the routine of intensive care units. The alarms of equipment intended to protect patients have increased noise within the unit, the level of distraction and interruptions in the workflow, leading to a false sense of security.

  11. Clinical Alarms in intensive care: implications of alarm fatigue for the safety of patients1

    Science.gov (United States)

    Bridi, Adriana Carla; Louro, Thiago Quinellato; da Silva, Roberto Carlos Lyra

    2014-01-01

    OBJECTIVES: to identify the number of electro-medical pieces of equipment in a coronary care unit, characterize their types, and analyze implications for the safety of patients from the perspective of alarm fatigue. METHOD: this quantitative, observational, descriptive, non-participatory study was conducted in a coronary care unit of a cardiology hospital with 170 beds. RESULTS: a total of 426 alarms were recorded in 40 hours of observation: 227 were triggered by multi-parametric monitors and 199 were triggered by other equipment (infusion pumps, dialysis pumps, mechanical ventilators, and intra-aortic balloons); that is an average of 10.6 alarms per hour. CONCLUSION: the results reinforce the importance of properly configuring physiological variables, the volume and parameters of alarms of multi-parametric monitors within the routine of intensive care units. The alarms of equipment intended to protect patients have increased noise within the unit, the level of distraction and interruptions in the workflow, leading to a false sense of security. PMID:25591100

  12. The heuristics of nurse responsiveness to critical patient monitor and ventilator alarms in a private room neonatal intensive care unit.

    Science.gov (United States)

    Joshi, Rohan; Mortel, Heidi van de; Feijs, Loe; Andriessen, Peter; Pul, Carola van

    2017-01-01

    Alarm fatigue is a well-recognized patient safety concern in intensive care settings. Decreased nurse responsiveness and slow response times to alarms are the potentially dangerous consequences of alarm fatigue. The aim of this study was to determine the factors that modulate nurse responsiveness to critical patient monitor and ventilator alarms in the context of a private room neonatal intensive care setting. The study design comprised of both a questionnaire and video monitoring of nurse-responsiveness to critical alarms. The Likert scale questionnaire, comprising of 50 questions across thematic clusters (critical alarms, yellow alarms, perception, design, nursing action, and context) was administered to 56 nurses (90% response rate). Nearly 6000 critical alarms were recorded from 10 infants in approximately 2400 hours of video monitoring. Logistic regression was used to identify patient and alarm-level factors that modulate nurse-responsiveness to critical alarms, with a response being defined as a nurse entering the patient's room within the 90s of the alarm being generated. Based on the questionnaire, the majority of nurses found critical alarms to be clinically relevant even though the alarms did not always mandate clinical action. Based on video observations, for a median of 34% (IQR, 20-52) of critical alarms, the nurse was already present in the room. For the remaining alarms, the response rate within 90s was 26%. The median response time was 55s (IQR, 37-70s). Desaturation alarms were the most prevalent and accounted for more than 50% of all alarms. The odds of responding to bradycardia alarms, compared to desaturation alarms, were 1.47 (95% CI = 1.21-1.78; heuristics in determining whether or not to respond to the alarm. Amongst other factors, the category and duration of critical alarms along with the clinical status of the patient determine nurse-responsiveness to alarms.

  13. ARC Code TI: Optimal Alarm System Design and Implementation

    Data.gov (United States)

    National Aeronautics and Space Administration — An optimal alarm system can robustly predict a level-crossing event that is specified over a fixed prediction horizon. The code contained in this packages provides...

  14. Central alarm system replacement in NPP Krsko

    International Nuclear Information System (INIS)

    Cicvaric, D.; Susnic, M.; Djetelic, N.

    2004-01-01

    Current NPP Krsko central alarm system consists of three main segments. Main Control Board alarm system (BETA 1000), Ventilation Control Board alarm system (BETA 1000) and Electrical Control Board alarm system (BETA 1100). All sections are equipped with specific BetaTone audible alarms and silence, acknowledge as well as test push buttons. The main reason for central alarm system replacement is system obsolescence and problems with maintenance, due to lack of spare parts. Other issue is lack of system redundancy, which could lead to loss of several Alarm Light Boxes in the event of particular power supply failure. Current central alarm system does not provide means of alarm optimization, grouping or prioritization. There are three main options for central alarm system replacement: Conventional annunciator system, hybrid annunciator system and advanced alarm system. Advanced alarm system implementation requires Main Control Board upgrade, integration of process instrumentation and plant process computer as well as long time for replacement. NPP Krsko has decided to implement hybrid alarm system with patchwork approach. The new central alarm system will be stand alone, digital, with advanced filtering and alarm grouping options. Sequence of event recorder will be linked with plant process computer and time synchronized with redundant GPS signal. Advanced functions such as link to plant procedures will be implemented with plant process computer upgrade in outage 2006. Central alarm system replacement is due in outage 2004.(author)

  15. Production of sensitivity and false alarm rate

    International Nuclear Information System (INIS)

    Han Zijie; Kang Wu; Chu Chengsheng; Hao Fanhua; Liu Xiaoya; Cao Lin; Hu Yongbo; Gong Jian; Xiang Yongchun; Zhang Jianhua; Yang Xiangdong

    2007-01-01

    The false alarm rate and sensitivity in nuclear material monitoring system are affected by alarm principle. Two different alarm principles are studied with theory and experiment analysis in this paper. Our research shows that theory analysis and experiment result are accordant very much. This study provides technology support for designing better alarm principle in nuclear material monitoring system. (authors)

  16. Reducing hospital noise: a review of medical device alarm management.

    Science.gov (United States)

    Konkani, Avinash; Oakley, Barbara; Bauld, Thomas J

    2012-01-01

    Increasing noise in hospital environments, especially in intensive care units (ICUs) and operating rooms (ORs), has created a formidable challenge for both patients and hospital staff. A major contributing factor for the increasing noise levels in these environments is the number of false alarms generated by medical devices. This study focuses on discovering best practices for reducing the number of false clinical alarms in order to increase patient safety and provide a quiet environment for both work and healing. The researchers reviewed Pub Med, Web of Knowledge and Google Scholar sources to obtain original journal research and review articles published through January 2012. This review includes 27 critically important journal articles that address different aspects of medical device alarms management, including the audibility, identification, urgency mapping, and response time of nursing staff and different solutions to such problems. With current technology, the easiest and most direct method for reducing false alarms is to individualize alarm settings for each patient's condition. Promoting an institutional culture change that emphasizes the importance of individualization of alarms is therefore an important goal. Future research should also focus on the development of smart alarms.

  17. Effects of RoHs and REACH regulations on firm-level production and export, and the role of global value chains : the cases of Malaysia and Vietnam

    OpenAIRE

    Otsuki, Tsunehiro; Honda, Keiichiro; Michida, Etsuyo; Nabeshima, Kaoru; Ueki, Yasushi

    2015-01-01

    This paper uses firm-level data to examine the impact of foreign chemical safety regulations such as RoHS and REACH on the production costs and export performance of firms in Malaysia and Vietnam. This paper also investigates the role of global value chains in enhancing the likelihood that a firm complies with RoHS and REACH. We find that in addition to the initial setup costs for compliance, EU RoHS (REACH) implementation imposes on firms additional variable production costs by requiring add...

  18. The Best Ever Alarm System Toolkit

    International Nuclear Information System (INIS)

    Kasemir, Kay; Chen, Xihui; Danilova, Ekaterina N.

    2009-01-01

    Learning from our experience with the standard Experimental Physics and Industrial Control System (EPICS) alarm handler (ALH) as well as a similar intermediate approach based on script-generated operator screens, we developed the Best Ever Alarm System Toolkit (BEAST). It is based on Java and Eclipse on the Control System Studio (CSS) platform, using a relational database (RDB) to store the configuration and log actions. It employs a Java Message Service (JMS) for communication between the modular pieces of the toolkit, which include an Alarm Server to maintain the current alarm state, an arbitrary number of Alarm Client user interfaces (GUI), and tools to annunciate alarms or log alarm related actions. Web reports allow us to monitor the alarm system performance and spot deficiencies in the alarm configuration. The Alarm Client GUI not only gives the end users various ways to view alarms in tree and table, but also makes it easy to access the guidance information, the related operator displays and other CSS tools. It also allows online configuration to be simply modified from the GUI. Coupled with a good 'alarm philosophy' on how to provide useful alarms, we can finally improve the configuration to achieve an effective alarm system.

  19. An Evidence-Based Approach to Reducing Cardiac Telemetry Alarm Fatigue.

    Science.gov (United States)

    Srinivasa, Ekta; Mankoo, Jaspreet; Kerr, Charles

    2017-08-01

    It is estimated that between 80% and 99% of alarms in the clinical areas are in actionable alarms (Gross, Dahl, & Nielson). Alarm management is one of the Joint Commission's National Patient Safety Goals (2014) because sentinel events have directly been linked to the devices generating these alarms. At an acute care facility in Boston, a multidisciplinary team consisting of Nursing, Biomedical Engineers, Patient Safety and Providers was formed to conduct a pilot study on the state of telemetry alarms on a surgical floor. An evidence-based approach was taken utilizing Philips Real-time data exporter alarms tracking software to capture all telemetry alarms during a 43-day time span. Likewise, noise meters were placed near telemetry alarm speakers to track decibel levels within the aforementioned timeframe for 21 days. Analysis of the data showed that clinically insignificant Premature Ventricular Contractions (PVC) alarms accounted for more than 40% of all alarms in the unit within the time span, while also contributing to an average noise level of 58.49 dB. In response to the data, the interdisciplinary team approved to permanently default the settings for PAIR PVC, MULTIFORM PVC, and RUN PVC alarms to off. The results showed a 54% decrease in the rate of alarms per bed per day, and an average noise reduction of 2.3 dB between the two selected noise measurement areas. Organizing a multidisciplinary team provides an effective framework toward analyzing and addressing cardiac telemetry alarm fatigue. Looking at quantitative datasets for clinical care areas through various lenses helps identify opportunities for improvement in regards to highlighting alarms that are not actionable. Pilot changes to alarm parameters can be tested for their environmental impact in the care area. © 2017 Sigma Theta Tau International.

  20. Nuthatches eavesdrop on variations in heterospecific chickadee mobbing alarm calls

    Science.gov (United States)

    Templeton, Christopher N.; Greene, Erick

    2007-01-01

    Many animals recognize the alarm calls produced by other species, but the amount of information they glean from these eavesdropped signals is unknown. We previously showed that black-capped chickadees (Poecile atricapillus) have a sophisticated alarm call system in which they encode complex information about the size and risk of potential predators in variations of a single type of mobbing alarm call. Here we show experimentally that red-breasted nuthatches (Sitta canadensis) respond appropriately to subtle variations of these heterospecific “chick-a-dee” alarm calls, thereby evidencing that they have gained important information about potential predators in their environment. This study demonstrates a previously unsuspected level of discrimination in intertaxon eavesdropping. PMID:17372225

  1. High construal level can help negotiators to reach integrative agreements: The role of information exchange and judgement accuracy.

    Science.gov (United States)

    Wening, Stefanie; Keith, Nina; Abele, Andrea E

    2016-06-01

    In negotiations, a focus on interests (why negotiators want something) is key to integrative agreements. Yet, many negotiators spontaneously focus on positions (what they want), with suboptimal outcomes. Our research applies construal-level theory to negotiations and proposes that a high construal level instigates a focus on interests during negotiations which, in turn, positively affects outcomes. In particular, we tested the notion that the effect of construal level on outcomes was mediated by information exchange and judgement accuracy. Finally, we expected the mere mode of presentation of task material to affect construal levels and manipulated construal levels using concrete versus abstract negotiation tasks. In two experiments, participants negotiated in dyads in either a high- or low-construal-level condition. In Study 1, high-construal-level dyads outperformed dyads in the low-construal-level condition; this main effect was mediated by information exchange. Study 2 replicated both the main and mediation effects using judgement accuracy as mediator and additionally yielded a positive effect of a high construal level on a second, more complex negotiation task. These results not only provide empirical evidence for the theoretically proposed link between construal levels and negotiation outcomes but also shed light on the processes underlying this effect. © 2015 The British Psychological Society.

  2. Object oriented design in the AGSDCS alarm system

    International Nuclear Information System (INIS)

    Morris, J.T.

    1994-01-01

    The alarm subsystem of the Alternating Gradient Synchrotron Distributed Control System (AGSDCS) has been redesigned to enhance reliability, flexibility, and ease of maintenance. The alarm system is functionally divided into Alarm Generators, Alarm Displays, and a central Alarm Receiver. The task was simplified by defining a set of C++ classes that could be reused by all components of the alarm system. The AlarmData class represents instances of alarm conditions. The AlarmFilter class is used by both the Alarm Receiver and Alarm Displays to select the alarms that are of interest to a particular user. The AlarmDatabase class is used by the Alarm Receiver to manage the central alarm database. The Alarm Displays use the AlarmDatabase class to manage the local database representing the alarms on their screens. ((orig.))

  3. Requirements for multi-level systems pharmacology models to reach end-usage : The case of type 2 diabetes

    NARCIS (Netherlands)

    Nyman, E.; Rozendaal, Y.J.W.; Helmlinger, G.; Hamrén, B.; Kjellsson, M.C.; Strålfors, P.; van Riel, N.A.W.; Gennemark, P.; Cedersund, G.

    2016-01-01

    We are currently in the middle of a major shift in biomedical research: unprecedented and rapidly growing amounts of data may be obtained today, from in vitro, in vivo and clinical studies, at molecular, physiological and clinical levels. To make use of these large-scale, multi-level datasets,

  4. Requirements for multi-level systems pharmacology models to reach end-usage: the case of type 2 diabetes

    NARCIS (Netherlands)

    Nyman, Elin; Rozendaal, Yvonne J. W.; Helmlinger, Gabriel; Hamrén, Bengt; Kjellsson, Maria C.; Strålfors, Peter; van Riel, Natal A. W.; Gennemark, Peter; Cedersund, Gunnar

    2016-01-01

    We are currently in the middle of a major shift in biomedical research: unprecedented and rapidly growing amounts of data may be obtained today, from in vitro, in vivo and clinical studies, at molecular, physiological and clinical levels. To make use of these large-scale, multi-level datasets,

  5. Approaching the Hard-to-Reach in Organized Colorectal Cancer Screening: an Overview of Individual, Provider and System Level Coping Strategies

    Directory of Open Access Journals (Sweden)

    Jason Liwen Huang

    2017-06-01

    Full Text Available Background: Despite the proven effectiveness of colorectal cancer (CRC screening on reduction of CRC mortality, the uptake of CRC screening remains low. Participation rate is one of determinants for the success of organized population-based screening program. This review aims to identify those who are hard-to-reach, and summarize the strategies to increase their screening rate from individual, provider and system levels. Methods: A systematic search of electronic English databases was conducted on the factors and strategies of uptake in CRC screening for the hard-to-reach population up to May 2017. Discussion: The coverage rate and participation rate are two indexes to identify the hard-to-reach population in organized CRC screening program. However, the homeless, new immigrants, people with severe mental illness, the jail intimates, and people with characteristics including lower education levels and/or low socioeconomic status, living in rural/remote areas, without insurance, and racial minorities are usually recognized as hard-to-reach populations. For them, organized screening programs offer a better coverage, while novel invitation approaches for eligible individuals and multiple strategies from primary care physicians are still needed to enhance screening rates among subjects who are hard-to-reach. Suggestions implied the effectiveness of interventions at the system level, including linkages to general practice; use of decision making tools; enlisting supports from coalition; and the continuum from screening to diagnosis and treatment. Conclusion: Organized CRC screening offers a system access to approach the hard-to-reach populations. To increase their uptake, multiple and novel strategies from individual, provider and system levels should be applied. For policymakers, public healthcare providers and community stakeholders, it is a test to tailor their potential needs and increase their participation rates through continuous efforts to

  6. False Alarms and Overmonitoring: Major Factors in Alarm Fatigue Among Labor Nurses.

    Science.gov (United States)

    Simpson, Kathleen Rice; Lyndon, Audrey

    2018-06-08

    Nurses can be exposed to hundreds of alarms during their shift, contributing to alarm fatigue. The purposes were to explore similarities and differences in perceptions of clinical alarms by labor nurses caring for generally healthy women compared with perceptions of adult intensive care unit (ICU) and neonatal ICU nurses caring for critically ill patients and to seek nurses' suggestions for potential improvements. Nurses were asked via focus groups about the utility of clinical alarms from medical devices. There was consensus that false alarms and too many devices generating alarms contributed to alarm fatigue, and most alarms lacked clinical relevance. Nurses identified certain types of alarms that they responded to immediately, but the vast majority of the alarms did not contribute to their clinical assessment or planned nursing care. Monitoring only those patients who need it and only those physiologic values that are warranted, based on patient condition, may decrease alarm burden.

  7. Hornbills can distinguish between primate alarm calls.

    Science.gov (United States)

    Rainey, Hugo J.; Zuberbühler, Klaus; Slater, Peter J. B.

    2004-01-01

    Some mammals distinguish between and respond appropriately to the alarm calls of other mammal and bird species. However, the ability of birds to distinguish between mammal alarm calls has not been investigated. Diana monkeys (Cercopithecus diana) produce different alarm calls to two predators: crowned eagles (Stephanoaetus coronatus) and leopards (Panthera pardus). Yellow-casqued hornbills (Ceratogymna elata) are vulnerable to predation by crowned eagles but are not preyed on by leopards and might therefore be expected to respond to the Diana monkey eagle alarm call but not to the leopard alarm call. We compared responses of hornbills to playback of eagle shrieks, leopard growls, Diana monkey eagle alarm calls and Diana monkey leopard alarm calls and found that they distinguished appropriately between the two predator vocalizations as well as between the two Diana monkey alarm calls. We discuss possible mechanisms leading to these responses. PMID:15209110

  8. Acculturation level and caregiver outcomes from a randomized intervention trial to enhance caregivers' health: evidence from REACH II.

    Science.gov (United States)

    Meyer, Oanh L; Liu, Xiaoyan Lucia; Tancredi, Daniel; Ramirez, A Susana; Schulz, Richard; Hinton, Ladson

    2018-06-01

    Latinos comprise a growing segment of the caregiver population and vary widely in acculturation, yet little is known regarding how acculturation might affect caregiver stress or intervention outcomes. This study examined the relationship between acculturation and burden, bother, and depression in Latino dementia caregivers at baseline and following an intervention. This was a secondary data analysis of 211 Latino caregivers of older adults with dementia from Resources for Enhancing Alzheimer's Caregiver Health (REACH) II, a multisite randomized trial of caregiver interventions. Baseline and follow-up data were used to run mixed-effects models examining the main and moderating effect of acculturation on caregiver stress. No significant main effect of acculturation was found for any of the outcome measures, controlling for demographic covariates. Acculturation moderated the effect of the intervention on caregiver burden: those who were more acculturated benefited more from the intervention. Differential acculturation for Latino caregivers was not directly associated with caregiver burden, bother, or depression, but was associated with reducing burden from the intervention. Future research should explore by what mechanism acculturation influences caregiver burden following an intervention.

  9. The perceived urgency of auditory warning alarms used in the hospital operating room is inappropriate.

    Science.gov (United States)

    Mondor, Todd A; Finley, G Allen

    2003-03-01

    To examine the perceived urgency of 13 auditory warning alarms commonly occurring in the hospital operating room. Undergraduate students, who were naïve with respect to the clinical situation associated with the alarms, judged perceived urgency of each alarm on a ten-point scale. The perceived urgency of the alarms was not consistent with the actual urgency of the clinical situation that triggers it. In addition, those alarms indicating patient condition were generally perceived as less urgent than those alarms indicating the operation of equipment. Of particular interest were three sets of alarms designed by equipment manufacturers to indicate specific priorities for action. Listeners did not perceive any differences in the urgency of the 'information only', 'medium' and 'high' priority alarms of two of the monitors with all judged as low to moderate in urgency. In contrast, the high priority alarm of the third monitor was judged as significantly more urgent than its low and medium urgency counterparts. The alarms currently in use do not convey the intended sense of urgency to naïve listeners, and this holds even for two sets of alarms designed specifically by manufacturers to convey different levels of urgency.

  10. Fire auto alarm system intelligent trend

    International Nuclear Information System (INIS)

    Du Chengbao

    1997-01-01

    The author gives the course and trend of the fire alarm system going to more computerized and more intelligent. It is described that only the system applied artificial intelligent and confusion control is the true intelligent fire alarm system. The author gives the detailed analysis on the signal treatment of artificial intelligent applied to analogue fire alarm system as well as the alarm system controlled by confusion technology and artificial nervous net

  11. Systems analysis of a security alarm system

    International Nuclear Information System (INIS)

    Schiff, A.

    1975-01-01

    When the Lawrence Livermore Laboratory found that its security alarm system was causing more false alarms and maintenance costs than LLL felt was tolerable, a systems analysis was undertaken to determine what should be done about the situation. This report contains an analysis of security alarm systems in general and ends with a review of the existing Security Alarm Control Console (SACC) and recommendations for its improvement, growth and change. (U.S.)

  12. 46 CFR 76.15-30 - Alarms.

    Science.gov (United States)

    2010-10-01

    ... than paint and lamp lockers and similar small spaces, shall be fitted with an approved audible alarm in... required to be fitted with a delayed discharge. Such alarms shall be so arranged as to sound during the 20 second delay period prior to the discharge of carbon dioxide into the space, and the alarm shall depend...

  13. Patient characteristics associated with false arrhythmia alarms in intensive care

    Directory of Open Access Journals (Sweden)

    Harris PR

    2017-04-01

    ventricular bradycardia as true or false. Total monitoring time for each patient was measured, and the number of false alarms per hour was calculated for these six alarm types. Medical records were examined to acquire data on patient characteristics. Results: A total of 461 unique patients (mean age =60±17 years were enrolled, generating a total of 2,558,760 alarms, including all levels of arrhythmia, parameter, and technical alarms. There were 48,404 hours of patient monitoring time, and an average overall alarm rate of 52 alarms/hour. Investigators annotated 12,671 arrhythmia alarms; 11,345 (89.5% were determined to be false. Two hundred and fifty patients (54% generated at least one of the six annotated alarm types. Two patients generated 6,940 arrhythmia alarms (55%. The number of false alarms per monitored hour for patients’ annotated arrhythmia alarms ranged from 0.0 to 7.7, and the duration of these false alarms per hour ranged from 0.0 to 158.8 seconds. Patient characteristics were compared in relation to 1 the number and 2 the duration of false arrhythmia alarms per 24-hour period, using nonparametric statistics to minimize the influence of outliers. Among the significant associations were the following: age ≥60 years (P=0.013; P=0.034, confused mental status (P<0.001 for both comparisons, cardiovascular diagnoses (P<0.001 for both comparisons, electrocardiographic (ECG features, such as wide ECG waveforms that correspond to ventricular depolarization known as QRS complex due to bundle branch block (BBB (P=0.003; P=0.004 or ventricular paced rhythm (P=0.002 for both comparisons, respiratory diagnoses (P=0.004 for both comparisons, and support with mechanical ventilation, including those with primary diagnoses other than respiratory ones (P<0.001 for both comparisons. Conclusion: Patients likely to trigger a higher number of false arrhythmia alarms may be those with older age, confusion, cardiovascular diagnoses, and ECG features that indicate BBB or

  14. Synergistic Success: The Power of Mentorship and Reaching the Senior Levels of the U.S. Army

    Science.gov (United States)

    2017-06-09

    2011, 50). Combined with the loss of manufacturing jobs in inner cities, the black community viewed attaining a skill to remain competitive in the... communication and team-building skills improving organizational performance. Barriers to Effective Mentorship Effective mentorship takes trust and time...level (National Center for Veterans Analysis and Statistics 2017). Creating value and building self - esteem will influence African American officers

  15. Reaching the hard to reach.

    Science.gov (United States)

    Bhiwandi, P; Campbell, M; Potts, M

    1994-01-01

    The 1994 International Conference on Population and Development proposed increasing contraceptive couple protection from 550 million in 1995 to 880 million in 2015. The task for family planning (FP) programs is to provide access to services for, sometimes, inaccessible rural populations. FP need based on desire for no more children has ranged from under 20% in Senegal to almost 80% in Peru. Socioeconomic development was found not to be a prerequisite for fertility change. Gender inequalities in education and social autonomy must be changed. FP access is very important among women with a disadvantaged background or among women unsure about FP. Bangladesh is a good example of a country with increased contraceptive prevalence despite low income. The rule of thumb is that contraception increases of 15% contribute to a drop in family size of about one child. Program effectiveness is related to a variety of factors: contraceptive availability at many locations, acceptable price of contraception, delivery of the oral contraceptives without prescriptions, and other strategies. FP is a service not a medical treatment. A range of methods must be promoted and available from a range of facilities. Contraceptive use is dependent on the woman's stage in her lifecycle and is dependent on informed choice. Community-based distribution systems are effective, whereas free distribution by poorly-trained field workers is not always very effective because patient payment of all or part of the cost assures quality and freedom of choice. Effective programs for underprivileged groups involve aggressive, easy to manage programs that can be replicated rapidly. FP serves a useful function in depressing maternal mortality among the poor in Africa, who have no access to quality health services. Social marketing is an effective strategy for reaching remote areas. Political will and robust management are necessary commodities.

  16. SUBSURFACE VISUAL ALARM SYSTEM ANALYSIS

    International Nuclear Information System (INIS)

    D.W. Markman

    2001-01-01

    The ''Subsurface Fire Hazard Analysis'' (CRWMS M andO 1998, page 61), and the document, ''Title III Evaluation Report for the Surface and Subsurface Communication System'', (CRWMS M andO 1999a, pages 21 and 23), both indicate the installed communication system is adequate to support Exploratory Studies Facility (ESF) activities with the exception of the mine phone system for emergency notification purposes. They recommend the installation of a visual alarm system to supplement the page/party phone system The purpose of this analysis is to identify data communication highway design approaches, and provide justification for the selected or recommended alternatives for the data communication of the subsurface visual alarm system. This analysis is being prepared to document a basis for the design selection of the data communication method. This analysis will briefly describe existing data or voice communication or monitoring systems within the ESF, and look at how these may be revised or adapted to support the needed data highway of the subsurface visual alarm. system. The existing PLC communication system installed in subsurface is providing data communication for alcove No.5 ventilation fans, south portal ventilation fans, bulkhead doors and generator monitoring system. It is given that the data communication of the subsurface visual alarm system will be a digital based system. It is also given that it is most feasible to take advantage of existing systems and equipment and not consider an entirely new data communication system design and installation. The scope and primary objectives of this analysis are to: (1) Briefly review and describe existing available data communication highways or systems within the ESF. (2) Examine technical characteristics of an existing system to disqualify a design alternative is paramount in minimizing the number of and depth of a system review. (3) Apply general engineering design practices or criteria such as relative cost, and degree

  17. HYBRID ALARM SYSTEMS: COMBINING SPATIAL ALARMS AND ALARM LISTS FOR OPTIMIZED CONTROL ROOM OPERATION

    Energy Technology Data Exchange (ETDEWEB)

    Ronald L. Boring; J.J. Persensky

    2012-07-01

    The US Department of Energy (DOE) is sponsoring research, development, and deployment on Light Water Reactor Sustainability (LWRS), in which the Idaho National Laboratory (INL) is working closely with nuclear utilities to develop technologies and solutions to help ensure the safe operational life extension of current nuclear power plants. One of the main areas of focus is control room modernization. Within control room modernization, alarm system upgrades present opportunities to meet the broader goals of the LWRS project in demonstrating the use and safety of the advanced instrumentation and control (I&C) technologies and the short-term and longer term objectives of the plant. In this paper, we review approaches for and human factors issues behind upgrading alarms in the main control room of nuclear power plants.

  18. [Level at which control objectives are reached in patients in different population groups with type 2 diabetes].

    Science.gov (United States)

    Herrero, A; Pinillos, J; Sabio, P; Martín, J L; Garzón, G; Gil, Á

    There is evidence of increased macro- and micro-vascular risk in diabetic patients. The objective of this study was to determine the level of control in patients in different population groups with type 2 diabetes. Descriptive cross-sectional study. Primary care. Madrid Health Service. Year: 2014. Patients over 14 years with type 2 diabetes. Number of patientes: n=6674. Variables on the degree of control (HbA1c, systolic blood pressure [SBP], diastolic blood pressure [DBP], LDL-c) and variables on patient characteristics (demographic, other cardiovascular risk factors, complications). The mean age of patients with controlled HbA1c was 67.8 years vs. 62.9 years in the uncontrolled (Pdifferences were statistically significant (P 140mmHg or DBP> 90mmHg. Over 25% of patients with hypertension or DL and uncontrolled levels were not receiving drug treatment. Control was improved in all groups, especially in younger patients, with particularly high cardiovascular risk by the presence of other cardiovascular risk factors or macroangiopathy. A significant percentage of patients with uncontrolled BP and cLDL were not diagnosed or receiving drug treatment. Copyright © 2016 Sociedad Española de Médicos de Atención Primaria (SEMERGEN). Publicado por Elsevier España, S.L.U. All rights reserved.

  19. A questionnaire comparison of two alarm systems

    International Nuclear Information System (INIS)

    Collier, Steven G.

    1997-11-01

    A questionnaire was developed, based on guidelines for alarm system design given in NUREG/CR-6105. The intentions were both to develop a subjective instrument for rating the effectiveness of alarm systems and to learn lessons on alarm system design from a comparison of two systems. The questionnaire was administered to reactor operations staff at two locations with different alarm systems embedded in a simulation of the same underlying PWR power plant: Loviisa NPP and Halden Man-Machine Laboratory. The questionnaire, considered as a measuring instrument, had good to high reliability and moderate to good content validity. The questionnaire is considered suitable for further use in the shortened form resulting from this study. Further work is also recommended. The degree of reliability and validity also lend a degree of validation to the NUREG guidelines. The questionnaire was able to show differences between ratings of the two alarm systems. The Loviisa system showed more consistency with other control room features and was better at drawing the operators' attention to important alarms. Both systems were not rated particularly well on alarm prioritisation and spurious alarms. The Halden system was better at showing naturally occurring relationships between alarms. Some of these differences may have been due to the subjects' greater familiarity with the Loviisa alarm system. The results nevertheless show that the questionnaire can measure subjective responses to alarm systems. (author)

  20. Evaluating alternative responses to safeguards alarms

    International Nuclear Information System (INIS)

    Al-Ayat, R.A.; Judd, B.R.; McCord, R.K.

    1982-01-01

    This paper describes a quantitative approach to help evaluate and respond to safeguards alarms. These alarms may be generated internally by a facility's safeguards systems or externally by individuals claiming to have stolen special nuclear material (SNM). This approach can be used to identify the most likely cause of an alarm - theft, hoax, or error - and to evaluate alternative responses to alarms. Possible responses include conducting investigations, initiating measures to recover stolen SNM, and replying to external threats. Based on the results of each alarm investigation step, the evaluation revises the likelihoods of possible causes of an alarm, and uses this information to determine the optimal sequence of further responses. The choice of an optimal sequence of responses takes into consideration the costs and benefits of successful thefts or hoaxes. These results provide an analytical basis for setting priorities and developing contingency plans for responding to safeguards alarms

  1. Video systems for alarm assessment

    International Nuclear Information System (INIS)

    Greenwoll, D.A.; Matter, J.C.; Ebel, P.E.

    1991-09-01

    The purpose of this NUREG is to present technical information that should be useful to NRC licensees in designing closed-circuit television systems for video alarm assessment. There is a section on each of the major components in a video system: camera, lens, lighting, transmission, synchronization, switcher, monitor, and recorder. Each section includes information on component selection, procurement, installation, test, and maintenance. Considerations for system integration of the components are contained in each section. System emphasis is focused on perimeter intrusion detection and assessment systems. A glossary of video terms is included. 13 figs., 9 tabs

  2. Video systems for alarm assessment

    Energy Technology Data Exchange (ETDEWEB)

    Greenwoll, D.A.; Matter, J.C. (Sandia National Labs., Albuquerque, NM (United States)); Ebel, P.E. (BE, Inc., Barnwell, SC (United States))

    1991-09-01

    The purpose of this NUREG is to present technical information that should be useful to NRC licensees in designing closed-circuit television systems for video alarm assessment. There is a section on each of the major components in a video system: camera, lens, lighting, transmission, synchronization, switcher, monitor, and recorder. Each section includes information on component selection, procurement, installation, test, and maintenance. Considerations for system integration of the components are contained in each section. System emphasis is focused on perimeter intrusion detection and assessment systems. A glossary of video terms is included. 13 figs., 9 tabs.

  3. An alarm multiplexer communication system

    International Nuclear Information System (INIS)

    Herrera, G.V.

    1986-01-01

    A low cost Alarm Multiplexer Communication System (AMCS) has been developed to perform the security sensor monitoring and control functions and to provide remote relay control capability for integrated security systems. AMCS has a distributed multiplexer/repeater architecture with up to four dual communication loops and dual control computers that guarantee total system operation under any single point failure condition. Each AMCS can control up to 4096 sensors and 2048 remote relays. AMCS reports alarm status information to and is controlled by either one or two Host computers. This allows for independent operation of primary and backup security command centers. AMCS communicates with the Host computers over an asynchronous serial communication link and has a message protocol which allows AMCS to fully recover from lost messages or large blocks of data communication errors. This paper describes the AMCS theory of operation, AMCS fault modes, and AMCS system design methodology. Also, cost and timing information is presented. AMCS is being used and considered for several DOE and DOD facilities

  4. Ecological alarm system for Itaipu

    Energy Technology Data Exchange (ETDEWEB)

    Faehser, L.

    1984-05-01

    At Itaipu, on the Rio Parana, Brazil and Paraguay are constructing the world's largest hydro-electric power plant with a capacity seven times as high as that of Assuan. An information system is intended to give fair warning in case of threatening ecological conditions. The computer-supported alarm system had four objectives: 1. presentation of the present ecological situation; 2. evaluation of the ecological risks; 3. warning about ecological deficits; 4. suggestions for establishing ecological stability. In a first step the available inventory data concerning soil, topography, vegetation and water were evaluated by expert groups according to their risk grade (0-4) and ecological weight (1-10). The product of these evaluations indicates the ecological deficit (0-40). At a threshold value of 30, the information system automatically signals ecological alarm and locates the centre of danger via computer-plotted maps and tables. The necessary data are supplied periodically by selected measurement stations. Quantification of ecological facts enables the persons who are responsible for decisions at Itaipu to recognize, avoid, or diminish elements of danger even if they have little or no ecological knowledge. The file of data that has been compiled so far should be extended parallel with the development in the Itaipu area. With the help of factor analysis connections of cause and effect can be detected in this extremely complex reservoir system which has hardly been explored yet.

  5. Receivers matter: the meaning of alarm calls and competition for nest sites in a bird community.

    Science.gov (United States)

    Parejo, Deseada; Avilés, Jesús M; Expósito-Granados, Mónica

    2018-04-11

    Animal communities may constitute information networks where individuals gain information on predation risk by eavesdropping on alarm calls of other species. However, communities include species in different trophic levels, and it is not yet known how the trophic level of the receiver influences the informative value of a call. Furthermore, no empirical study has yet tested how increased competition may influence the value of alarm calls for distinct receivers. Here, we identify the importance of alarm calls emitted by a small owl, the little owl (Athene noctua), on the structure of a cavity-nesting bird community including mesopredators and primary prey under variable levels of competition for nest holes. Competitors sharing top predators with the callers and prey of the callers interpreted alarm and non-alarm calls differently. Competitors chose preferentially alarm and non-alarm patches over control patches to breed, while prey selected alarm patches. In contrast, competition for nest sites affected habitat selection of prey species more than that of competitors of the callers. This study provides support for a changing value of alarm calls and competition for nest sites for distinct receivers related to niche overlapping among callers and eavesdroppers, therefore, calling attention to possible cascading effects by the use of information in natural communities.

  6. Auditory backup alarms: distance-at-first-detection via in-situ experimentation on alarm design and hearing protection effects.

    Science.gov (United States)

    Alali, Khaled; Casali, John G

    2012-01-01

    The purpose of this study was to assess normal hearing listeners' performance in detecting a stationary backup alarm signal and to quantify the linear distance at detection point. Detection distances for 12 participants with normal hearing were measured while they were fitted with 7 hearing protectors and while they were unoccluded (open ear). A standard (narrowband) backup alarm signal and a broadband (pulsed white noise) backup alarm signal from Brigade[1] were used. The method of limits, with distance as the physical measurement variable and threshold detection as the task, was employed to find at which distance the participant could first detect the backup alarms. A within-subject Analysis of Variance (ANOVA) revealed a significant main effect of the listening conditions on the detection distance in feet. Post hoc analyses indicated that the Bilsom L3HV conventional passive earmuff (at 1132.2 ft detection distance) was significantly poorer compared to all other HPDs and the open ear in detection distance achieved, and that there were no statistically-significant differences between the unoccluded ear (1652.3 ft), EB-15-Lo BlastPLGTM (1546.2 ft), EB-15-Hi BlastPLGTM (1543.4 ft), E-A-R/3M Combat ArmsTM earplug-nonlinear, level-dependent state (1507.8 ft), E-A-R/3M HiFiTM earplug (1497.7 ft), and Bilsom ImpactTM dichotic electronic earmuff (1567.2 ft). In addition, the E-A-R/3M Combat ArmsTM earplug-passive steady state resulted in significantly longer detection distances than only the open ear condition, at 1474.1 ft versus 1652.3 ft for the open ear. ANOVA also revealed a significant main effect of the backup alarm type on detection distance. The means were 1600.9 ft for the standard (narrowband) backup alarm signal, and a significantly closer 1379.4 ft was required for the Brigade broadband backup alarm signal. For on-ground workers, it is crucial to detect backup alarm signals as far away as possible rather than at close distances since this will provide them

  7. Urethral alarm probe for permanent prostate implants

    International Nuclear Information System (INIS)

    Cutajar, D.; Lerch, M.; Takacs, G.

    2008-01-01

    We have developed a urethral dosimetry system for real time dose verification along the urethra during permanent implant prostate brachytherapy. The urethral alarm uses 'spectroscopic dosimetry' to calculate the dose rate along the urethra in real time. The application of spectroscopic dosimetry for the urethral alarm probe was verified using Monte Carlo calculations. In phantom depth dose measurements as well as isotropy measurements were performed to verify the usefulness of the urethra alarm probe as an in vivo real time dosimeter. (author)

  8. Reducing false asystole alarms in intensive care.

    Science.gov (United States)

    Dekimpe, Remi; Heldt, Thomas

    2017-07-01

    High rates of false monitoring alarms in intensive care can desensitize staff and therefore pose a significant risk to patient safety. Like other critical arrhythmia alarms, asystole alarms require immediate attention by the care providers as a true asystole event can be acutely life threatening. Here, it is illustrated that most false asystole alarms can be attributed to poor signal quality, and we propose and evaluate an algorithm to identify data windows of poor signal quality and thereby help suppress false asystole alarms. The algorithm combines intuitive signal-quality features (degree of signal saturation and baseline wander) and information from other physiological signals that might be available. Algorithm training and testing was performed on the MIMIC II and 2015 PhysioNet/Computing in Cardiology Challenge databases, respectively. The algorithm achieved an alarm specificity of 81.0% and sensitivity of 95.4%, missing only one out of 22 true asystole alarms. On a separate neonatal data set, the algorithm was able to reject 89.7% (890 out of 992) of false asystole alarms while keeping all 22 true events. The results show that the false asystole alarm rate can be significantly reduced through basic signal quality evaluation.

  9. Identification and Isolation of Human Alarm Pheromones

    National Research Council Canada - National Science Library

    Mujica-Parodi, Lilianne R; Strey, Helmut

    2006-01-01

    .... Task I, Optimization of Sample Collection, focused on the collection of the putative alarm pheromone via axillary sweat samples obtained during reference (physical exercise) and arousal (skydive) conditions...

  10. Advanced alarm systems: Display and processing issues

    Energy Technology Data Exchange (ETDEWEB)

    O`Hara, J.M. [Brookhaven National Lab., Upton, NY (United States); Wachtel, J.; Perensky, J. [US Nuclear Regulatory Commission, Washington, DC (United States). Office of Nuclear Regulatory Research

    1995-05-01

    This paper describes a research program sponsored by the US Nuclear Regulatory Commission to address the human factors engineering (HFE) deficiencies associated with nuclear power plant alarm systems. The overall objective of the study is to develop HFE review guidance for alarm systems. In support of this objective, human performance issues needing additional research were identified. Among the important issues were alarm processing strategies and alarm display techniques. This paper will discuss these issues and briefly describe our current research plan to address them.

  11. Comparison of sound propagation and perception of three types of backup alarms with regards to worker safety

    Directory of Open Access Journals (Sweden)

    Véronique Vaillancourt

    2013-01-01

    Full Text Available A technology of backup alarms based on the use of a broadband signal has recently gained popularity in many countries. In this study, the performance of this broadband technology is compared to that of a conventional tonal alarm and a multi-tone alarm from a worker-safety standpoint. Field measurements of sound pressure level patterns behind heavy vehicles were performed in real work environments and psychoacoustic measurements (sound detection thresholds, equal loudness, perceived urgency and sound localization were carried out in the laboratory with human subjects. Compared with the conventional tonal alarm, the broadband alarm generates a much more uniform sound field behind vehicles, is easier to localize in space and is judged slighter louder at representative alarm levels. Slight advantages were found with the tonal alarm for sound detection and for perceived urgency at low levels, but these benefits observed in laboratory conditions would not overcome the detrimental effects associated with the large and abrupt variations in sound pressure levels (up to 15-20 dB within short distances observed in the field behind vehicles for this alarm, which are significantly higher than those obtained with the broadband alarm. Performance with the multi-tone alarm generally fell between that of the tonal and broadband alarms on most measures.

  12. The effectiveness of nurse education and training for clinical alarm response and management: a systematic review.

    Science.gov (United States)

    Yue, Liqing; Plummer, Virginia; Cross, Wendy

    2017-09-01

    To identify the effectiveness of education interventions provided for nurses for clinical alarm response and management. Some education has been undertaken to improve clinical alarm response, but the evidence for evaluating effectiveness for nurse education interventions is limited. Systematic review. A systematic review of experimental studies published in English from 2005-2015 was conducted in four computerised databases (MEDLINE, EMBASE, CINAHL and Scopus). After identification, screening and appraisal using Joanna Briggs Institute instruments, quality research papers were selected, data extraction and analysis followed. Five studies met the inclusion criteria for alarm response and no articles were concerned with clinical alarm education for management. All had different types and methods of interventions and statistical pooling was not possible. Response accuracy, response time and perceptions were consistent when different interventions were adopted. A positive effect was identified when learning about general alarms, single alarms, sequential alarms and medium-level alarms for learning as the primary task. Nurses who were musically trained had a faster and more accurate alarm response. Simulation interventions had a positive effect, but the effect of education provided in the care unit was greater. Overall, clinical alarm awareness was improved through education activities. Nurses are the main users of healthcare alarms and work in complex environments with high numbers of alarms, including nuisance alarms and other factors. Alarm-related adverse events are common. The findings of a small number of experimental studies with diverse evidence included consideration of various factors when formulating education strategies. The factors which influence effectiveness of nurse education are nurse demographics, nurse participants with musical training, workload and characteristics of alarms. Education interventions based in clinical practice settings increase

  13. The impact of recent changes in smoke alarm legislation on residential fire injuries and smoke alarm ownership in New South Wales, Australia.

    Science.gov (United States)

    Harvey, Lara A; Poulos, Roslyn G; Sherker, Shauna

    2013-01-01

    In 2006, New South Wales (NSW) state legislation changed from requiring smoke alarms in new houses only to all houses. We evaluated the impact of this legislative change on residential fire injury and smoke alarm ownership characteristics. Residential fire injuries for 2002 to 2010 were identified from hospitalization data for all hospitals in NSW. Data relating to smoke alarm ownership and demographic factors were obtained from the NSW Population Health Survey. Negative binomial regression analysis was used to analyze trends over time. Prior to the introduction of universal legislation, hospitalization rates were increasing slightly; however, following the introduction of legislation, hospitalization rates decreased by an estimated 36.2% (95% confidence interval [CI], 16.7-55.8) annually. Smoke alarm ownership increased from 73.3% (95% CI, 72.5-74.2) prelegislation to 93.6% (95% CI, 93.1-94.2) 18 months postlegislation. Thirty percent of households reported testing their alarms regularly. Speaking a language other than English (relative risks [RRs], 1.82; 95% CI, 1.44-2.99), allowing smoking in the home (RR, 1.73; 95% CI, 1.31-2.27), and being part of the most disadvantaged socioeconomic group (RR, 1.47; 95% CI, 1.14-1.91) remain major risk factors for nonownership. Broadening the scope of state legislation has had a positive impact on residential fire-related hospitalizations and smoke alarm ownership. However, it is of concern that the legislation has been the least effective in increasing smoke alarm ownership among non-English-speaking households, in households where smoking is allowed, in low socioeconomic households, and that a high proportion of householders do not test their smoke alarms regularly. Targeted campaigns are needed to reach these high-risk groups and to ensure that smoke alarms are functional.

  14. Preclinical evidence of mitochondrial nicotinamide adenine dinucleotide as an effective alarm parameter under hypoxia

    Science.gov (United States)

    Shi, Hua; Sun, Nannan; Mayevsky, Avraham; Zhang, Zhihong; Luo, Qingming

    2014-01-01

    Early detection of tissue hypoxia in the intensive care unit is essential for effective treatment. Reduced nicotinamide adenine dinucleotide (NADH) has been suggested to be the most sensitive indicator of tissue oxygenation at the mitochondrial level. However, no experimental evidence comparing the kinetics of changes in NADH and other physiological parameters has been provided. The aim of this study is to obtain the missing data in a systematic and reliable manner. We constructed four acute hypoxia models, including hypoxic hypoxia, hypemic hypoxia, circulatory hypoxia, and histogenous hypoxia, and measured NADH fluorescence, tissue reflectance, cerebral blood flow, respiration, and electrocardiography simultaneously from the induction of hypoxia until death. We found that NADH was not always the first onset parameter responding to hypoxia. The order of responses was mainly affected by the cause of hypoxia. However, NADH reached its alarm level earlier than the other monitored parameters, ranging from several seconds to >10 min. As such, we suggest that the NADH can be used as a hypoxia indicator, although the exact level that should be used must be further investigated. When the NADH alarm is detected, the body still has a chance to recover if appropriate and timely treatment is provided.

  15. Gamin partable radiation meter with alarm threshold

    International Nuclear Information System (INIS)

    Payat, Rene.

    1981-10-01

    The Gamin Radiation meter is a direct reading, portable, battery-powered gamma doserate meter featuring alarm thresholds. Doserate is read on a micro-ammeter with a millirad-per-hour logarithmic scale, covering a range of 0,1 to 1000 millirads/hour. The instrument issues an audible warning signal when dose-rate level exceeds a threshold value, which can be selected. The detector tube is of the Geiger-Muller counter, energy compensated type. Because of its low battery drain, the instrument can be operated continously for 1000 hours. It is powered by four 1.5 volt alcaline batteries of the R6 type. The electronic circuitry is housed in a small lightweight case made of impact resistant plastic. Applications of the Gamin portable radiation monitor are found in health physics, safety departments, medical facilities, teaching, civil defense [fr

  16. Effects of integrated designs of alarm and process information on diagnosis performance in digital nuclear power plants.

    Science.gov (United States)

    Wu, Xiaojun; She, Manrong; Li, Zhizhong; Song, Fei; Sang, Wei

    2017-12-01

    In the main control rooms of nuclear power plants (NPPs), operators frequently switch between alarm displays and system-information displays to incorporate information from different screens. In this study, we investigated two integrated designs of alarm and process information - integrating alarm information into process displays (denoted as Alarm2Process integration) and integrating process information into alarm displays (denoted as Process2Alarm integration). To analyse the effects of the two integration approaches and time pressure on the diagnosis performance, a laboratory experiment was conducted with ninety-six students. The results show that compared with the non-integrated case, Process2Alarm integration yields better diagnosis performance in terms of diagnosis accuracy, time required to generate correct hypothesis and completion time. In contrast, the Alarm2Process integration leads to higher levels of workload, with no improvement in diagnosis performance. The diagnosis performance of Process2Alarm integration was consistently better than that of Alarm2Process integration, regardless of the levels of time pressure. Practitioner Summary: To facilitate operator's synthesis of NPP information when performing diagnosis tasks, we proposed to integrate process information into alarm displays. The laboratory validation shows that the integration approach significantly improves the diagnosis performance for both low and high time-pressure levels.

  17. The CANDU alarm analysis tool (CAAT)

    Energy Technology Data Exchange (ETDEWEB)

    Davey, E C; Feher, M P; Lupton, L R [Control Centre Technology Branch, ON (Canada)

    1997-09-01

    AECL undertook the development of a software tool to assist alarm system designers and maintainers based on feedback from several utilities and design groups. The software application is called the CANDU Alarm Analysis Tool (CAAT) and is being developed to: Reduce by one half the effort required to initially implement and commission alarm system improvements; improve the operational relevance, consistency and accuracy of station alarm information; record the basis for alarm-related decisions; provide printed reports of the current alarm configuration; and, make day-to-day maintenance of the alarm database less tedious and more cost-effective. The CAAT assists users in accessing, sorting and recording relevant information, design rules, decisions, and provides reports in support of alarm system maintenance, analysis of design changes, or regulatory inquiry. The paper discusses the need for such a tool, outlines the application objectives and principles used to guide tool development, describes the how specific tool features support user design and maintenance tasks, and relates the lessons learned from early application experience. (author). 4 refs, 2 figs.

  18. T-Farm complex alarm upgrades

    Energy Technology Data Exchange (ETDEWEB)

    Roberts, J.B.

    1995-01-01

    The alarm and controls associated with the T, TX, and TY farms are located in the 242-T control room. The design data for replacement and upgrades of the alarm panels is in this document. This task was canceled previous to the 90% design review point.

  19. The CANDU alarm analysis tool (CAAT)

    International Nuclear Information System (INIS)

    Davey, E.C.; Feher, M.P.; Lupton, L.R.

    1997-01-01

    AECL undertook the development of a software tool to assist alarm system designers and maintainers based on feedback from several utilities and design groups. The software application is called the CANDU Alarm Analysis Tool (CAAT) and is being developed to: Reduce by one half the effort required to initially implement and commission alarm system improvements; improve the operational relevance, consistency and accuracy of station alarm information; record the basis for alarm-related decisions; provide printed reports of the current alarm configuration; and, make day-to-day maintenance of the alarm database less tedious and more cost-effective. The CAAT assists users in accessing, sorting and recording relevant information, design rules, decisions, and provides reports in support of alarm system maintenance, analysis of design changes, or regulatory inquiry. The paper discusses the need for such a tool, outlines the application objectives and principles used to guide tool development, describes the how specific tool features support user design and maintenance tasks, and relates the lessons learned from early application experience. (author). 4 refs, 2 figs

  20. Useful and usable alarm systems : recommended properties

    International Nuclear Information System (INIS)

    Veland, Oeystein; Kaarstad, Magnhild; Seim, Lars Aage; Foerdestroemmen, Nils T.

    2001-01-01

    This document describes the result of a study on alarm systems conducted by IFE in Halden. The study was initiated by the Norwegian Petroleum Directorate. The objective was to identify and formulate a set of important properties for useful and usable alarm systems. The study is mainly based on review of the latest international recognised guidelines and standards on alarm systems available at the time of writing, with focus on realistic solutions from research and best practice from different industries. In addition, IFE experiences gathered through specification and design of alarm systems and experimental activities in HAMMLAB and bilateral projects, have been utilized where relevant. The document presents a total of 43 recommendations divided into a number of general recommendations and more detailed recommendations on alarm generation, structuring, prioritisation, presentation and handling. (Author)

  1. Alarm processing system using AI techniques for nuclear power plant

    International Nuclear Information System (INIS)

    Yang, Joon On; Chang, Soon Heung

    1990-01-01

    An alarm processing system (APS) has been developed using artificial intelligence (AI) techniques. The alarms of nuclear power plants (NPP's) are classified into the generalized and special alarms. The generalized alarms are also classified into the global and local alarms. For each type of alarms, the specific processing rules are applied to filter and suppress unnecessary and potentially misleading alarms. The local processing are based on 'model-based reasoning.' The global and special alarms are processed by using the general cause-consequence check rules. The priorities of alarms are determined according to the plant state and the consistencies between them

  2. Comprehensive smoke alarm coverage in lower economic status homes: alarm presence, functionality, and placement.

    Science.gov (United States)

    Sidman, Elanor A; Grossman, David C; Mueller, Beth A

    2011-08-01

    The objectives of this study are to estimate smoke alarm coverage and adherence with national guidelines in low- to mid-value owner-occupied residences, and to identify resident demographic, behavioral, and building characteristics and other fire and burn safety practices associated with smoke alarm utilization. Baseline visits were conducted with 779 households in King County, Washington, for a randomized trial of smoke alarm functionality. Presence, functionality, features, and location of pre-existing smoke alarms were ascertained by staff observation and testing. Household and building descriptors were collected using questionnaires. Households were classified by presence of smoke alarms, functional alarms, and functional and properly mounted alarms placed in hallways and on each floor but not in recommended avoidance locations. Smoke alarms were present in 89%, and functional units in 78%, of households. Only 6-38% met all assessed functionality and placement recommendations. Homes frequently lacked alarms in any bedrooms or on each floor. Building age, but not renovation status, was associated with all dimensions of smoke alarm coverage; post-1980 constructions were 1.7 times more likely to comply with placement recommendations than were pre-1941 homes (95% CI: 1.1-2.6). Respondent education and race/ethnicity, children wood stoves and fireplaces, number of smoke alarms, recency of smoke alarm testing, carbon monoxide monitors, and fire ladders displayed varying relationships with alarm presence, functionality, and placement. Strategies for maintaining smoke alarms in functional condition and improving compliance with placement recommendations are necessary to achieve universal coverage, and will benefit the majority of households.

  3. Reactor alarm system development and application issues

    Energy Technology Data Exchange (ETDEWEB)

    Drexler, J E; Oicese, G O [INVAP S.E. (Argentina)

    1997-09-01

    The new hardware and software technologies, and the need in research reactors for assistance systems in operation and maintenance, have given an appropriate background to develop a computer based system named ``Reactor Alarm System`` (RAS). RAS is a software package, user oriented, with emphasis on production, experiments and maintenance goals. It is designed to run on distributed systems conformed with microcomputers under QNX operating system. RAS main features are: (a) Alarm Panel Display; (b) Alarm Page; (c) Alarm Masking and Inhibition; (d) Alarms Color and Attributes; (e) Condition Classification; and (f) Arrangement Presentation. RAS design allows it to be installed as a part of a computer based Supervision and Control System in new installations or retrofit existing reactor instrumentation systems. The analysis of human factors during development stage and successive user feedback from different applications, brought out several RAS improvements: (a) Multiple-copy alarm summaries; (b) Improved alarm handling; (c) Extended dictionary; and (d) Enhanced hardware availability. It has proved successful in providing new capabilities for operators, and also has shown the continuous increase of user-demands, reflecting the expectations placed today on computer-based systems. (author). 6 figs, 1 tabs.

  4. Reactor alarm system development and application issues

    International Nuclear Information System (INIS)

    Drexler, J.E.; Oicese, G.O.

    1997-01-01

    The new hardware and software technologies, and the need in research reactors for assistance systems in operation and maintenance, have given an appropriate background to develop a computer based system named ''Reactor Alarm System'' (RAS). RAS is a software package, user oriented, with emphasis on production, experiments and maintenance goals. It is designed to run on distributed systems conformed with microcomputers under QNX operating system. RAS main features are: a) Alarm Panel Display; b) Alarm Page; c) Alarm Masking and Inhibition; d) Alarms Color and Attributes; e) Condition Classification; and f) Arrangement Presentation. RAS design allows it to be installed as a part of a computer based Supervision and Control System in new installations or retrofit existing reactor instrumentation systems. The analysis of human factors during development stage and successive user feedback from different applications, brought out several RAS improvements: a) Multiple-copy alarm summaries; b) Improved alarm handling; c) Extended dictionary; and d) Enhanced hardware availability. It has proved successful in providing new capabilities for operators, and also has shown the continuous increase of user-demands, reflecting the expectations placed today on computer-based systems. (author). 6 figs, 1 tabs

  5. When one is not enough: prevalence and characteristics of homes not adequately protected by smoke alarms.

    Science.gov (United States)

    Peek-Asa, C; Allareddy, V; Yang, J; Taylor, C; Lundell, J; Zwerling, C

    2005-12-01

    The National Fire Protection Association (NFPA) has specific recommendations about the number, location, and type of smoke alarms that are needed to provide maximum protection for a household. No previous studies have examined whether or not homes are completely protected according to these guidelines. The authors describe the prevalence and home characteristics associated with compliance to recommendations for smoke alarm installation by the NFPA. Data are from the baseline on-site survey of a randomized trial to measure smoke alarm effectiveness. The trial was housed in a longitudinal cohort study in a rural Iowa county. Of 1005 homes invited, 691 (68.8%) participated. Information about smoke alarm type, placement, and function, as well as home and occupant characteristics, was collected through an on-site household survey. Although 86.0% of homes had at least one smoke alarm, only 22.3% of homes (approximately one in five) were adequately protected according to NFPA guidelines. Fourteen percent of homes had no functioning smoke alarms. More than half of the homes with smoke alarms did not have enough of them or had installed them incorrectly, and 42.4% of homes with alarms had at least one alarm that did not operate. Homes with at least one high school graduate were nearly four times more likely to be fully protected. Homes that had multiple levels, a basement, or were cluttered or poorly cleaned were significantly less likely to be fully protected. These findings indicate that consumers may not be knowledgeable about the number of alarms they need or how to properly install them. Occupants are also not adequately maintaining the alarms that are installed.

  6. Alarm points for fixed oxygen monitors

    International Nuclear Information System (INIS)

    Miller, G.C.

    1987-05-01

    Oxygen concentration monitors were installed in a vault where numerous pipes carried inert cryogens and gases to the Mirror Fusion Test Facility (MFTF-B) experimental vessel at Lawrence Livermore National Laboratory (LLNL). The problems associated with oxygen-monitoring systems and the reasons why such monitors were installed were reviewed. As a result of this review, the MFTF-B monitors were set to sound an evacuation alarm when the oxygen concentration fell below 18%. We chose the 18% alarm criterion to minimize false alarms and to allow time for personnel to escape in an oxygen-deficient environment

  7. Alarm criteria in radiation portal monitoring

    Energy Technology Data Exchange (ETDEWEB)

    Burr, Tom [Los Alamos National Laboratory, Statistical Sciences Group, Mail Stop F600, Los Alamos, NM 87545 (United States)]. E-mail: tburr@lanl.gov; Gattiker, James R. [Los Alamos National Laboratory, Statistical Sciences Group, Mail Stop F600, Los Alamos, NM 87545 (United States); Myers, Kary [Los Alamos National Laboratory, Statistical Sciences Group, Mail Stop F600, Los Alamos, NM 87545 (United States); Tompkins, George [Los Alamos National Laboratory, Risk Analysis and Decision Support Systems, Stop F609, Los Alamos, NM 87545 (United States)

    2007-05-15

    Gamma detectors at border crossings are intended to detect illicit nuclear material. These detectors collect counts that are used to determine whether to trigger an alarm. Several candidate alarm rules are evaluated, with attention to background suppression caused by the vehicle. Because the count criterion leads to many nuisance alarms and because background suppression by the vehicle is smaller for ratios of counts, analysis of a ratio criterion is included. Detection probability results that consider the effects of 5 factors are given for 2 signal-injection studies, 1 for counts, and 1 for count ratios.

  8. Alarm criteria in radiation portal monitoring

    International Nuclear Information System (INIS)

    Burr, Tom; Gattiker, James R.; Myers, Kary; Tompkins, George

    2007-01-01

    Gamma detectors at border crossings are intended to detect illicit nuclear material. These detectors collect counts that are used to determine whether to trigger an alarm. Several candidate alarm rules are evaluated, with attention to background suppression caused by the vehicle. Because the count criterion leads to many nuisance alarms and because background suppression by the vehicle is smaller for ratios of counts, analysis of a ratio criterion is included. Detection probability results that consider the effects of 5 factors are given for 2 signal-injection studies, 1 for counts, and 1 for count ratios

  9. Wallac automatic alarm dosimeter type RAD21

    International Nuclear Information System (INIS)

    Burgess, P. H.; Iles, W.J.

    1980-02-01

    The Automatic Alarm Dosimeter type RAD 21 is a batterypowered personal dosemeter and exposure rate alarm monitor, designed to be worn on the body, covering an exposure range from 0.1 to 999.9 mR and has an audible alarm which can be pre-set over the range 1 mR h -1 to 250 mR h -1 . The instrument is designed to measure x- and γ radiation over the energy range 50 keV to 3 MeV. The facilities and controls, the radiation, electrical, environmental and mechanical characteristics, and the manual, have been evaluated. (U.K.)

  10. Percutaneous nephrolithotomy; alarming variables for postoperative bleeding

    Directory of Open Access Journals (Sweden)

    Shakhawan H.A. Said

    2017-03-01

    Conclusion: According to our present results stone complexity (GSS grade 3 and 4, history of ipsilateral renal stone surgery, and occurrence of intraoperative pelvicalyceal perforation are alarming variables for post-PCNL bleeding.

  11. The nature of alarm communication in Constrictotermes cyphergaster (Blattodea: Termitoidea: Termitidae: the integration of chemical and vibroacoustic signals

    Directory of Open Access Journals (Sweden)

    Paulo F. Cristaldo

    2015-12-01

    Full Text Available Alarm signalling is of paramount importance to communication in all social insects. In termites, vibroacoustic and chemical alarm signalling are bound to operate synergistically but have never been studied simultaneously in a single species. Here, we inspected the functional significance of both communication channels in Constrictotermes cyphergaster (Termitidae: Nasutitermitinae, confirming the hypothesis that these are not exclusive, but rather complementary processes. In natural situations, the alarm predominantly attracts soldiers, which actively search for the source of a disturbance. Laboratory testing revealed that the frontal gland of soldiers produces a rich mixture of terpenoid compounds including an alarm pheromone. Extensive testing led to identification of the alarm pheromone being composed of abundant monoterpene hydrocarbons (1S-α-pinene and myrcene, along with a minor component, (E-β-ocimene. The vibratory alarm signalling consists of vibratory movements evidenced as bursts; a series of beats produced predominantly by soldiers. Exposing termite groups to various mixtures containing the alarm pheromone (crushed soldier heads, frontal gland extracts, mixture of all monoterpenes, and the alarm pheromone mixture made of standards resulted in significantly higher activity in the tested groups and also increased intensity of the vibratory alarm communication, with the responses clearly dose-dependent. Lower doses of the pheromone provoked higher numbers of vibratory signals compared to higher doses. Higher doses induced long-term running of all termites without stops necessary to perform vibratory behaviour. Surprisingly, even crushed worker heads led to low (but significant increases in the alarm responses, suggesting that other unknown compound in the worker's head is perceived and answered by termites. Our results demonstrate the existence of different alarm levels in termites, with lower levels being communicated through

  12. Addressing the alarm analysis barrier - a tool for improving alarm systems

    Energy Technology Data Exchange (ETDEWEB)

    Davey, E C; Basso, R A; Feher, M P [Atomic Energy of Canada Ltd., Chalk River, ON (Canada)

    1996-12-31

    This paper describes a software application tool for the initial specification and maintenance of the thousands of alarms in nuclear and other process control plants. The software program is used by system designers and maintainers to analyze, characterize, record and maintain the alarm information and configuration decisions for an alarm system. The tool provides a comprehensive design and information handling environment for: the existing alarm functions in current CANDU plants; the new alarm processing and presentation concepts developed under CANDU Owners Group (COG) sponsorship that are available to be applied to existing CANDU plants on a retrofit basis; and, the alarm functions to be implemented in new CANDU plants. (author). 3 refs., 1 fig.

  13. Addressing the alarm analysis barrier - a tool for improving alarm systems

    International Nuclear Information System (INIS)

    Davey, E.C.; Basso, R.A.; Feher, M.P.

    1995-01-01

    This paper describes a software application tool for the initial specification and maintenance of the thousands of alarms in nuclear and other process control plants. The software program is used by system designers and maintainers to analyze, characterize, record and maintain the alarm information and configuration decisions for an alarm system. The tool provides a comprehensive design and information handling environment for: the existing alarm functions in current CANDU plants; the new alarm processing and presentation concepts developed under CANDU Owners Group (COG) sponsorship that are available to be applied to existing CANDU plants on a retrofit basis; and, the alarm functions to be implemented in new CANDU plants. (author). 3 refs., 1 fig

  14. Sensor fusion for intelligent alarm analysis

    International Nuclear Information System (INIS)

    Nelson, C.L.; Fitzgerald, D.S.

    1996-01-01

    The purpose of an intelligent alarm analysis system is to provide complete and manageable information to a central alarm station operator by applying alarm processing and fusion techniques to sensor information. This paper discusses the sensor fusion approach taken to perform intelligent alarm analysis for the Advanced Exterior Sensor (AES). The AES is an intrusion detection and assessment system designed for wide-area coverage, quick deployment, low false/nuisance alarm operation, and immediate visual assessment. It combines three sensor technologies (visible, infrared, and millimeter wave radar) collocated on a compact and portable remote sensor module. The remote sensor module rotates at a rate of 1 revolution per second to detect and track motion and provide assessment in a continuous 360 degree field-of-regard. Sensor fusion techniques are used to correlate and integrate the track data from these three sensors into a single track for operator observation. Additional inputs to the fusion process include environmental data, knowledge of sensor performance under certain weather conditions, sensor priority, and recent operator feedback. A confidence value is assigned to the track as a result of the fusion process. This helps to reduce nuisance alarms and to increase operator confidence in the system while reducing the workload of the operator

  15. Judging the urgency of non-verbal auditory alarms: a case study.

    Science.gov (United States)

    Arrabito, G Robert; Mondor, Todd; Kent, Kimberley

    2004-06-22

    When designed correctly, non-verbal auditory alarms can convey different levels of urgency to the aircrew, and thereby permit the operator to establish the appropriate level of priority to address the alarmed condition. The conveyed level of urgency of five non-verbal auditory alarms presently used in the Canadian Forces CH-146 Griffon helicopter was investigated. Pilots of the CH-146 Griffon helicopter and non-pilots rated the perceived urgency of the signals using a rating scale. The pilots also ranked the urgency of the alarms in a post-experiment questionnaire to reflect their assessment of the actual situation that triggers the alarms. The results of this investigation revealed that participants' ratings of perceived urgency appear to be based on the acoustic properties of the alarms which are known to affect the listener's perceived level of urgency. Although for 28% of the pilots the mapping of perceived urgency to the urgency of their perception of the triggering situation was statistically significant for three of the five alarms, the overall data suggest that the triggering situations are not adequately conveyed by the acoustic parameters inherent in the alarms. The pilots' judgement of the triggering situation was intended as a means of evaluating the reliability of the alerting system. These data will subsequently be discussed with respect to proposed enhancements in alerting systems as it relates to addressing the problem of phase of flight. These results call for more serious consideration of incorporating situational awareness in the design and assignment of auditory alarms in aircraft.

  16. 46 CFR 97.37-9 - Carbon dioxide alarm.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 4 2010-10-01 2010-10-01 false Carbon dioxide alarm. 97.37-9 Section 97.37-9 Shipping... Markings for Fire and Emergency Equipment, Etc. § 97.37-9 Carbon dioxide alarm. (a) All carbon dioxide alarms shall be conspicuously identified: “WHEN ALARM SOUNDS—VACATE AT ONCE. CARBON DIOXIDE BEING...

  17. 46 CFR 78.47-9 - Carbon dioxide alarm.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 3 2010-10-01 2010-10-01 false Carbon dioxide alarm. 78.47-9 Section 78.47-9 Shipping... and Emergency Equipment, Etc. § 78.47-9 Carbon dioxide alarm. (a) All carbon dioxide alarms shall be conspicuously identified: “WHEN ALARM SOUNDS—VACATE AT ONCE. CARBON DIOXIDE BEING RELEASED.” (b) [Reserved] ...

  18. 46 CFR 169.732 - Carbon dioxide alarm.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 7 2010-10-01 2010-10-01 false Carbon dioxide alarm. 169.732 Section 169.732 Shipping... Control, Miscellaneous Systems, and Equipment Markings § 169.732 Carbon dioxide alarm. Each carbon dioxide alarm must be conspicuously identified: “WHEN ALARM SOUNDS—VACATE AT ONCE. CARBON DIOXIDE BEING RELEASED.” ...

  19. 46 CFR 196.37-9 - Carbon dioxide alarm.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 7 2010-10-01 2010-10-01 false Carbon dioxide alarm. 196.37-9 Section 196.37-9 Shipping... Markings for Fire and Emergency Equipment, etc. § 196.37-9 Carbon dioxide alarm. (a) All carbon dioxide alarms shall be conspicuously identified: “WHEN ALARM SOUNDS—VACATE AT ONCE. CARBON DIOXIDE BEING...

  20. 46 CFR 108.627 - Carbon dioxide alarm.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 4 2010-10-01 2010-10-01 false Carbon dioxide alarm. 108.627 Section 108.627 Shipping... EQUIPMENT Equipment Markings and Instructions § 108.627 Carbon dioxide alarm. Each carbon dioxide alarm must be identified by marking: “WHEN ALARM SOUNDS VACATE AT ONCE. CARBON DIOXIDE BEING RELEASED” next to...

  1. An evaluation approach for alarm processing improvement

    International Nuclear Information System (INIS)

    Kim, Jung-Taek; Lee, Dong-Young; Hwang, In-Koo; Park, Jae-Chang

    1997-01-01

    In light of the need to improve MMIS of NPPs, the advanced I and C research team of KAERI has embarked on developing an Alarm and Diagnosis-Integrated Operator Support System, called ADIOS, to filter or suppress unnecessary or nuisance alarms and diagnose abnormality of the plant process. ADIOS has been built in an object-oriented AI environment of G-2 expert system software tool, as presented in a companion paper. ADIOS then is evaluated according to the plan in three steps; (1) preliminary tests to refine the knowledge base and inference structure of ADIOS in such a dynamic environment, and also to evaluate the appropriateness of alarm-processing algorithms; (2) to ensure correctness, consistency, and completeness in the knowledge base using COKEP (Checker Of Knowledge base using Extended Petri net); and (3) the cognitive performance evaluation using the Simulation Analyzer with a Cognitive Operator Model (SACOM) in the KAERI's Integrated Test Facility (ITF). (author). 5 figs, 1 tab

  2. Behavioral alarm treatment for nocturnal enuresis

    Directory of Open Access Journals (Sweden)

    Rodrigo F. Pereira

    2010-06-01

    Full Text Available PURPOSES: To investigate the efficacy of alarm treatment in a sample of Brazilian children and adolescents with nocturnal enuresis and relate treatment success to age and type of clinical support. MATERIALS AND METHODS: During 32 weeks, 84 children and adolescents received alarm treatment together with weekly psychological support sessions for individual families or groups of 5 to 10 families. RESULTS: 71% of the participants achieved success, defined as 14 consecutive dry nights. The result was similar for children and adolescents and for individual or group support. The time until success was shorter for participants missing fewer support sessions. CONCLUSIONS: Alarm treatment was effective for the present sample, regardless of age or type of support. Missing a higher number of support sessions, which may reflect low motivation for treatment, increased the risk of failure.

  3. System for alarms analysis and optimization in petrochemicals plants; Sistema para analise e otimizacao de alarmes em plantas petroquimicas

    Energy Technology Data Exchange (ETDEWEB)

    Leitao, Gustavo; Pifer, Aderson; Guedes, Luiz Affonso [Universidade Federal do Rio Grande do Norte (UFRN), Natal, RN (Brazil); Saito, Kaku; Aquino, Leonardo [PETROBRAS, Rio de Janeiro, RJ (Brazil). Centro de Pesquisas (CENPES)

    2008-07-01

    The present work presents a group of algorithms, techniques and functionalities on alarms management which can be used efficiently on the treatment of 'disturbances' caused by the informal management of the alarm systems. Among the disturbances handled by these techniques, there is the recognition of intermittent alarms and false alarms, location of alarm floods and correlation between alarms, aiming the identification of communal root causes. The results will be presented through a case study on petrochemical alarm plants. At last, the results obtained by the utilization of such functionalities will be presented and discussed. (author)

  4. Nuisance alarm suppression techniques for fibre-optic intrusion detection systems

    Science.gov (United States)

    Mahmoud, Seedahmed S.; Visagathilagar, Yuvaraja; Katsifolis, Jim

    2012-02-01

    The suppression of nuisance alarms without degrading sensitivity in fibre-optic intrusion detection systems is important for maintaining acceptable performance. Signal processing algorithms that maintain the POD and minimize nuisance alarms are crucial for achieving this. A level crossings algorithm is presented for suppressing torrential rain-induced nuisance alarms in a fibre-optic fence-based perimeter intrusion detection system. Results show that rain-induced nuisance alarms can be suppressed for rainfall rates in excess of 100 mm/hr, and intrusion events can be detected simultaneously during rain periods. The use of a level crossing based detection and novel classification algorithm is also presented demonstrating the suppression of nuisance events and discrimination of nuisance and intrusion events in a buried pipeline fibre-optic intrusion detection system. The sensor employed for both types of systems is a distributed bidirectional fibre-optic Mach Zehnder interferometer.

  5. The alarm system of the SAPHIR detector

    International Nuclear Information System (INIS)

    Schultz-Coulon, H.C.

    1993-06-01

    In order to obtain an effective control of the different detector components an alarm system was built and implemented into the data acquisition system of the SAPHIR experiment. It provides an easy way of indicating errors by either adequate library calls or an appropriate hardware signal, both leading to an active alarm. This allows to react directly to any error detected by one of the specific control systems. In addition for selected kinds of errors the data run can be stopped automatically. Concept and construction of this system are described and some examples for its application are given. (orig.)

  6. Coincidence logic modules for criticality alarming

    International Nuclear Information System (INIS)

    Schaief, C.C. III.

    1977-04-01

    A coincidence Logic Module and a companion contact closure Relay Module utilizing the NIM Standard have been developed for criticality alarming. The units provide an ALARM whenever two or more out of N detectors become activated. In addition, an ALERT is generated whenever one or more detectors is activated or when certain electronic component failures occur. The number of detector inputs (N) can be expanded in groups of six by adding modules. Serial and parallel redundancy were used to reduce the probability of system failure

  7. Myocardial infarction false alarm: initial electrocardiogram and cardiac enzymes.

    Science.gov (United States)

    Gupta, Esha Das; Sakthiswary, Rajalingham

    2014-05-01

    The objectives of this study were to determine the incidence of a myocardial infarction "false alarm" and evaluate the efficacy of the initial electrocardiogram and cardiac enzymes in diagnosing myocardial infarction in Malaysia. We recruited patients who were admitted with suspected myocardial infarction from June to August 2008. The medical records of these patients were reviewed for the initial electrocardiogram, initial cardiac enzyme levels (creatinine kinase-MB and troponin T), and the final diagnosis upon discharge. The subjects were stratified into 2 groups: true myocardial infarction, and false alarm. 125 patients were enrolled in this study. Following admission and further evaluation, the diagnosis was revised from myocardial infarction to other medical conditions in 48 (38.4%) patients. The sensitivity and specificity of the initial ischemic electrocardiographic changes were 54.5% and 70.8%, respectively. Raised cardiac enzymes had a sensitivity of 44.3% and specificity of 95.8%. A significant proportion of patients in Malaysia are admitted with a false-alarm myocardial infarction. The efficacy of the electrocardiogram in diagnosing myocardial infarction in Malaysia was comparable to the findings of Western studies, but the cardiac enzymes had a much lower sensitivity.

  8. Seismic alarm system for Ignalina nuclear power plant

    International Nuclear Information System (INIS)

    Wieland, M.; Griesser, L.; Austin, G.E.; Tiurin, S.; Kuendig, C.

    2001-01-01

    A seismic alarm system will be installed at the Ignalina Nuclear Power Plant (INPP) in Lithuania. There are two reactors, both RMBK 1500 MW units. Each reactor is a water cooled, graphite moderated, channel type reactor. INPP has the most advanced version of the RMBK reactor design series. The first and second units of INPP went into service at the end of 1983 and in August 1987 respectively. Their design lifetime is approx. 30 years. The various buildings and plant have been designed for two earthquake levels, that is the design earthquake and the maximum possible earthquake with peak ground accelerations ranging from 1.2% to 10% of the acceleration due to gravity. Certain parts of the buildings and some of the equipment of the first and second units do not comply with Western seismic standards. As seismic strengthening of the existing buildings and equipment is not feasible economically, a reactor protection system based on an earthquake early warning system was recommended. This system essentially consists of six seismic stations encircling INPP at a radial distance of approx. 30 km and a seventh station at INPP. Each station includes three seismic substations each 500 m apart. The ground motion at each station is measured continuously by three accelerometers and one seismometer. Data is transmitted via telemetry to the control centre at INPP. Early warning alarms are generated if a seismic threshold is exceeded. This paper discusses the characteristics of INPP, the seismic alarm system presently under construction and the experience with other early warning and seismic alarm systems. (author)

  9. A revival of the alarm system: Making the alarm list useful during incidents

    International Nuclear Information System (INIS)

    Larsson, J. E.; Oehman, B.; Calzada, A.; Nihlwing, C.; Jokstad, H.; Kristianssen, L. I.; Kvalem, J.; Lind, M.

    2006-01-01

    In control rooms there are often problems with information overload, which means that the operators may receive more information than they are able to interpret. The most serious information overload occurs in two types of situations. The first is when the operating state of the plant changes, which often gives raise to a shower of alarms and events. Such an alarm shower is expected, but can be dangerous, because it may hide other alarms originating from unrelated faults. The second problem occurs when a fault causes several consequential faults, leading to a so-called alarm cascade. Because the alarms seldom arrive in correct time order, it can be very difficult to analyze such a cascade, and the information overload occurs in exactly the moment when a potentially dangerous situation starts. In an ongoing project, GoalArt and IFE are cooperating in testing and evaluating GoalArt's methods for alarm reduction and root cause analysis. The testing comprises two specific algorithms, root cause analysis and state-based alarm priority. The GoalArt system has been integrated with the HAMBO simulator so that operators can evaluate the algorithms on-line. (authors)

  10. The ecology and evolution of avian alarm call signaling systems

    Science.gov (United States)

    Billings, Alexis Chandon

    Communication is often set up as a simple dyadic exchange between one sender and one receiver. However, in reality, signaling systems have evolved and are used with many forms and types of information bombarding multiple senders, who in turn send multiple signals of different modalities, through various environmental spaces, finally reaching multiple receivers. In order to understand both the ecology and evolution of a signaling system, we must examine all the facets of the signaling system. My dissertation focused on the alarm call signaling system in birds. Alarm calls are acoustic signals given in response to danger or predators. My first two chapters examine how information about predators alters alarm calls. In chapter one I found that chickadees make distinctions between predators of different hunting strategies and appear to encode information about predators differently if they are heard instead of seen. In my second chapter, I test these findings more robustly in a non-model bird, the Steller's jay. I again found that predator species matters, but that how Steller's jays respond if they saw or heard the predator depends on the predator species. In my third chapter, I tested how habitat has influenced the evolution of mobbing call acoustic structure. I found that habitat is not a major contributor to the variation in acoustic structure seen across species and that other selective pressures such as body size may be more important. In my fourth chapter I present a new framework to understand the evolution of multimodal communication across species. I identify a unique constraint, the need for overlapping sensory systems, thresholds and cognitive abilities between sender and receiver in order for different forms of interspecific communication to evolve. Taken together, these chapters attempt to understand a signaling system from both an ecological and evolutionary perspective by examining each piece of the communication scheme.

  11. Development of the newly advanced alarm system for APWR plant

    Energy Technology Data Exchange (ETDEWEB)

    Shimada, Manabu; Yamamoto, Yoshihiro; Tani, Mamoru; Kobashi, Shuichi [Kansai Electric Power Co., Inc., Osaka (Japan)

    1997-09-01

    We have been developing AMCB (Advanced Main Control Board) for APWR consisting of a large overview display and on operator console. We have adopted the alarm prioritizing functions, which are already in use in the existing Japanese PWR plants, for easier identification of the high priority alarms. Moreover, we have developed an alarm system with a large overview display, which presents alarms on the plant process flow diagram. This enhances the location aids and pattern recognition in the alarm identification process. This time, we made further improvement and studies for better and various functions combining a large overview display with a CRT display. We determined the alarm system specification as follows, taking account of flexible alarm recognition processes. (1) The high priority alarms can be identified upon the LOD (large overview display). On the display, the alarms are described on the plant flow diagram, and the alarm status is shown on the fixed position of process or equipment symbols. (2) Other alarms are identified on large overview display and on CRTs using a hierarchical process. (3) The alarm messages are divided into 4 different groups according to the plant systems, thus enabling to undertake the countermeasure operations, using only the CRT. Moreover, we integrated a computerized ARPs (Alarm Response Procedures) into the alarm system. (author). 4 figs, 5 tabs.

  12. Development of the newly advanced alarm system for APWR plant

    International Nuclear Information System (INIS)

    Shimada, Manabu; Yamamoto, Yoshihiro; Tani, Mamoru; Kobashi, Shuichi

    1997-01-01

    We have been developing AMCB (Advanced Main Control Board) for APWR consisting of a large overview display and on operator console. We have adopted the alarm prioritizing functions, which are already in use in the existing Japanese PWR plants, for easier identification of the high priority alarms. Moreover, we have developed an alarm system with a large overview display, which presents alarms on the plant process flow diagram. This enhances the location aids and pattern recognition in the alarm identification process. This time, we made further improvement and studies for better and various functions combining a large overview display with a CRT display. We determined the alarm system specification as follows, taking account of flexible alarm recognition processes. (1) The high priority alarms can be identified upon the LOD (large overview display). On the display, the alarms are described on the plant flow diagram, and the alarm status is shown on the fixed position of process or equipment symbols. (2) Other alarms are identified on large overview display and on CRTs using a hierarchical process. (3) The alarm messages are divided into 4 different groups according to the plant systems, thus enabling to undertake the countermeasure operations, using only the CRT. Moreover, we integrated a computerized ARPs (Alarm Response Procedures) into the alarm system. (author). 4 figs, 5 tabs

  13. 46 CFR 95.15-30 - Alarms.

    Science.gov (United States)

    2010-10-01

    ... navigated, other than paint and lamp lockers and similar small spaces, shall be fitted with an approved... only for systems required to be fitted with a delayed discharge. Such alarms shall be so arranged as to sound during the 20 second delay period prior to the discharge of carbon dioxide into the space, and the...

  14. Adjustable electronic load-alarm relay

    International Nuclear Information System (INIS)

    Mason, C.H.; Sitton, R.S.

    1976-01-01

    An improved electronic alarm relay for monitoring the current drawn by an ac motor or other electrical load is described. The circuit is designed to measure the load with high accuracy and to have excellent alarm repeatability. Chattering and arcing of the relay contacts are minimal. The operator can adjust the set point easily and can re-set both the high and the low alarm points by means of one simple adjustment. The relay includes means for generating a signal voltage proportional to the motor current. In a preferred form of the invention a first operational amplifier is provided to generate a first constant reference voltage which is higher than a preselected value of the signal voltage. A second operational amplifier is provided to generate a second constant reference voltage which is lower than the aforementioned preselected value of the signal voltage. A circuit comprising a first resistor serially connected to a second resistor is connected across the outputs of the first and second amplifiers, and the junction of the two resistors is connected to the inverting terminal of the second amplifier. Means are provided to compare the aforementioned signal voltage with both the first and second reference voltages and to actuate an alarm if the signal voltage is higher than the first reference voltage or lower than the second reference voltage

  15. 10 CFR 74.57 - Alarm resolution.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 2 2010-01-01 2010-01-01 false Alarm resolution. 74.57 Section 74.57 Energy NUCLEAR REGULATORY COMMISSION (CONTINUED) MATERIAL CONTROL AND ACCOUNTING OF SPECIAL NUCLEAR MATERIAL Formula... unresolved beyond the time period specified for its resolution in the licensee's fundamental nuclear material...

  16. Object-oriented alarm-filtering system

    International Nuclear Information System (INIS)

    Corsberg, D.R.; Wilkie, D.

    1986-01-01

    This paper discusses an alarm-filtering system (AFS) being developed by EG and G Idaho, Inc. for the Advanced Test Reactor (ATR) at the Idaho National Engineering Laboratory. The ultimate goal of this project is to place AFS into ATR's reactor control room to act as an aid during major plant transients. In addition, methods of alarm analysis are investigated based on functional relationships rather than on a historical approach utilizing cause-consequence trees. Artificial intelligence techniques, including object-oriented programming, are also demonstrated as useful in analyzing alarms and alarm sequences. After a brief description of the problem AFS addresses, this paper discusses the design constraints and human factors that influenced the development of the system. The reader is then presented with operational and architectural descriptions of the system as well as what directions the future development of AFS may take. The fact that AFS is being considered as a partial solution to the problems discussed in the next section demonstrates the viability of its underlying technology and approach. 10 refs

  17. An Undergraduate Experiment in Alarm System Design.

    Science.gov (United States)

    Martini, R. A.; And Others

    1988-01-01

    Describes an experiment involving data acquisition by a computer, digital signal transmission from the computer to a digital logic circuit and signal interpretation by this circuit. The system is being used at the Illinois Institute of Technology. Discusses the fundamental concepts involved. Demonstrates the alarm experiment as it is used in…

  18. Reach Address Database (RAD)

    Data.gov (United States)

    U.S. Environmental Protection Agency — The Reach Address Database (RAD) stores the reach address of each Water Program feature that has been linked to the underlying surface water features (streams,...

  19. Nuclear power plant alarm systems: Problems and issues

    International Nuclear Information System (INIS)

    O'Hara, J.M.; Brown, W.S.

    1991-01-01

    Despite the incorporation of advanced technology into nuclear power plant alarm systems, human factors problems remain. This paper identifies to be addressed in order to allow advanced technology to be used effectively in the design of nuclear power plant alarm systems. The operator's use and processing of alarm system information will be considered. Based upon a review of alarm system research, issues related to general system design, alarm processing, display and control are discussed. It is concluded that the design of effective alarm systems depends on an understanding of the information processing capabilities and limitations of the operator. 39 refs

  20. Nuclear power plant alarm systems: Problems and issues

    Energy Technology Data Exchange (ETDEWEB)

    O' Hara, J.M.; Brown, W.S.

    1991-01-01

    Despite the incorporation of advanced technology into nuclear power plant alarm systems, human factors problems remain. This paper identifies to be addressed in order to allow advanced technology to be used effectively in the design of nuclear power plant alarm systems. The operator's use and processing of alarm system information will be considered. Based upon a review of alarm system research, issues related to general system design, alarm processing, display and control are discussed. It is concluded that the design of effective alarm systems depends on an understanding of the information processing capabilities and limitations of the operator. 39 refs.

  1. Perspectives on use of personal alarms by older fallers

    Directory of Open Access Journals (Sweden)

    Kylie Johnston

    2010-08-01

    Full Text Available Kylie Johnston1, Karen Grimmer-Somers1, Michele Sutherland21International Centre for Allied Health Evidence, University of South Australia, Adelaide; 2Falls Prevention Unit, Department of Health, Government of South Australia, Adelaide, AustraliaBackground: Personal alarms are proposed as a reliable mechanism for older people to obtain assistance after falling. However, little is known about how older people feel about owning and using personal alarms.Aim: This paper reports on experiences of independently living older people, who have recently fallen, regarding alarm use and their independence.Method: Volunteers older than 65 years who had sustained a fall in the previous six months were sought via community invitations. Semistructured telephone interviews were conducted to gain information about their fall and their perspectives on personal alarm use. Interviews were content-analyzed to identify key concepts and themes.Results: Thirty-one interviews were conducted. Twenty callers owned personal alarms. Four subgroups of older fallers were identified; the first group used personal alarms effectively and were advocates for their benefits, the second group owned an alarm but did not use it effectively, the third group did not own alarms mostly because of cost, although were receptive to an alarm should one be provided, and the fourth group did not have an alarm and would not use it even if it was provided.Discussion: Personal alarms produce positive experiences when used effectively by the right people. The cost of personal alarms prohibits some older fallers from being effective alarm users. However, other elderly fallers remain unwilling to consider alarm use even if one was provided. In view of their cost, personal alarms should be targeted to people who will benefit most. ­Alternative strategies should be considered when alarms are unlikely to be used appropriately.Keywords: personal alarm devices, falls, older people, patient perspective

  2. Reaching the hard-to-reach.

    Science.gov (United States)

    Valdes, C

    1992-01-01

    Guatemala's family planning (FP) programs are innovative but contraceptive use is only 23%. Total fertility is 5.3 children/woman, and the 9.5 million population will double in 23 years. The problem is poverty and illiteracy among rural residents removed from health services. 80% live in poverty and 80% are illiterate. Government effort is devoted to combating diseases such as diarrhea so there are few funds for implementing a comprehensive population policy. There is support within the national government but FP lacks priority status. APROFAM's goals are to use innovative marketing methods to inform the rural population who lack access to and knowledge about FP. Service delivery is constrained by the difficulty in reaching remote areas where 4 out of 10 indigenous Guatemalans live. Infant mortality can reach as high as 200/1000 live births. Population growth has slowed, and APROFAM plans to reach 16,000 more in the future. Promotions are conducted in several languages and aired on radio, television, and in the print media. It has been found that market research is the most effective strategy in reaching indigenous families. APROFAM has also been effective in upgrading service facilities through training, client surveys, and setting improved clinic standards. Breastfeeding, training, and voluntary sterilization programs contribute to the primary care effort. The example is given of Paulina Lebron from a very poor area who has learned how to space her children and thus improve the standard of living for her family. Eventually, she convinced herself and her family that sterilization was necessary, and now the couple enjoy the bliss of newlyweds without fear of pregnancy.

  3. Habituation of adult sea lamprey repeatedly exposed to damage-released alarm and predator cues

    Science.gov (United States)

    Imre, Istvan; Di Rocco, Richard T.; Brown, Grant E.; Johnson, Nicholas

    2016-01-01

    Predation is an unforgiving selective pressure affecting the life history, morphology and behaviour of prey organisms. Selection should favour organisms that have the ability to correctly assess the information content of alarm cues. This study investigated whether adult sea lamprey Petromyzon marinus habituate to conspecific damage-released alarm cues (fresh and decayed sea lamprey extract), a heterospecific damage-released alarm cue (white sucker Catostomus commersoniiextract), predator cues (Northern water snake Nerodia sipedon washing, human saliva and 2-phenylethylamine hydrochloride (PEA HCl)) and a conspecific damage-released alarm cue and predator cue combination (fresh sea lamprey extract and human saliva) after they were pre-exposed 4 times or 8 times, respectively, to a given stimulus the previous night. Consistent with our prediction, adult sea lamprey maintained an avoidance response to conspecific damage-released alarm cues (fresh and decayed sea lamprey extract), a predator cue presented at high relative concentration (PEA HCl) and a conspecific damage-released alarm cue and predator cue combination (fresh sea lamprey extract plus human saliva), irrespective of previous exposure level. As expected, adult sea lamprey habituated to a sympatric heterospecific damage-released alarm cue (white sucker extract) and a predator cue presented at lower relative concentration (human saliva). Adult sea lamprey did not show any avoidance of the Northern water snake washing and the Amazon sailfin catfish extract (heterospecific control). This study suggests that conspecific damage-released alarm cues and PEA HCl present the best options as natural repellents in an integrated management program aimed at controlling the abundance of sea lamprey in the Laurentian Great Lakes.

  4. False alarm reduction in BSN-based cardiac monitoring using signal quality and activity type information.

    Science.gov (United States)

    Tanantong, Tanatorn; Nantajeewarawat, Ekawit; Thiemjarus, Surapa

    2015-02-09

    False alarms in cardiac monitoring affect the quality of medical care, impacting on both patients and healthcare providers. In continuous cardiac monitoring using wireless Body Sensor Networks (BSNs), the quality of ECG signals can be deteriorated owing to several factors, e.g., noises, low battery power, and network transmission problems, often resulting in high false alarm rates. In addition, body movements occurring from activities of daily living (ADLs) can also create false alarms. This paper presents a two-phase framework for false arrhythmia alarm reduction in continuous cardiac monitoring, using signals from an ECG sensor and a 3D accelerometer. In the first phase, classification models constructed using machine learning algorithms are used for labeling input signals. ECG signals are labeled with heartbeat types and signal quality levels, while 3D acceleration signals are labeled with ADL types. In the second phase, a rule-based expert system is used for combining classification results in order to determine whether arrhythmia alarms should be accepted or suppressed. The proposed framework was validated on datasets acquired using BSNs and the MIT-BIH arrhythmia database. For the BSN dataset, acceleration and ECG signals were collected from 10 young and 10 elderly subjects while they were performing ADLs. The framework reduced the false alarm rate from 9.58% to 1.43% in our experimental study, showing that it can potentially assist physicians in diagnosing a vast amount of data acquired from wireless sensors and enhance the performance of continuous cardiac monitoring.

  5. False Alarm Reduction in BSN-Based Cardiac Monitoring Using Signal Quality and Activity Type Information

    Directory of Open Access Journals (Sweden)

    Tanatorn Tanantong

    2015-02-01

    Full Text Available False alarms in cardiac monitoring affect the quality of medical care, impacting on both patients and healthcare providers. In continuous cardiac monitoring using wireless Body Sensor Networks (BSNs, the quality of ECG signals can be deteriorated owing to several factors, e.g., noises, low battery power, and network transmission problems, often resulting in high false alarm rates. In addition, body movements occurring from activities of daily living (ADLs can also create false alarms. This paper presents a two-phase framework for false arrhythmia alarm reduction in continuous cardiac monitoring, using signals from an ECG sensor and a 3D accelerometer. In the first phase, classification models constructed using machine learning algorithms are used for labeling input signals. ECG signals are labeled with heartbeat types and signal quality levels, while 3D acceleration signals are labeled with ADL types. In the second phase, a rule-based expert system is used for combining classification results in order to determine whether arrhythmia alarms should be accepted or suppressed. The proposed framework was validated on datasets acquired using BSNs and the MIT-BIH arrhythmia database. For the BSN dataset, acceleration and ECG signals were collected from 10 young and 10 elderly subjects while they were performing ADLs. The framework reduced the false alarm rate from 9.58% to 1.43% in our experimental study, showing that it can potentially assist physicians in diagnosing a vast amount of data acquired from wireless sensors and enhance the performance of continuous cardiac monitoring.

  6. The false alarm at Forsmark March 6th 1986

    International Nuclear Information System (INIS)

    Hultaaker, Oe.

    1986-01-01

    118 households were interviewed by telephone. Additional interviews were made with the representatives of the nuclear plant, county alarming center, Radio Uppland and the authorities in question. Six out of ten households heard the false alarm. Less than one out of twenty believed that it was an alarm caused by real danger and seven out of ten think that there will be more false alarms. (G.B.)

  7. Automated Information System (AIS) Alarm System

    Energy Technology Data Exchange (ETDEWEB)

    Hunteman, W.

    1997-05-01

    The Automated Information Alarm System is a joint effort between Los Alamos National Laboratory, Lawrence Livermore National Laboratory, and Sandia National Laboratory to demonstrate and implement, on a small-to-medium sized local area network, an automated system that detects and automatically responds to attacks that use readily available tools and methodologies. The Alarm System will sense or detect, assess, and respond to suspicious activities that may be detrimental to information on the network or to continued operation of the network. The responses will allow stopping, isolating, or ejecting the suspicious activities. The number of sensors, the sensitivity of the sensors, the assessment criteria, and the desired responses may be set by the using organization to meet their local security policies.

  8. Automated Information System (AIS) Alarm System

    International Nuclear Information System (INIS)

    Hunteman, W.

    1997-01-01

    The Automated Information Alarm System is a joint effort between Los Alamos National Laboratory, Lawrence Livermore National Laboratory, and Sandia National Laboratory to demonstrate and implement, on a small-to-medium sized local area network, an automated system that detects and automatically responds to attacks that use readily available tools and methodologies. The Alarm System will sense or detect, assess, and respond to suspicious activities that may be detrimental to information on the network or to continued operation of the network. The responses will allow stopping, isolating, or ejecting the suspicious activities. The number of sensors, the sensitivity of the sensors, the assessment criteria, and the desired responses may be set by the using organization to meet their local security policies

  9. Technical guide to criticality alarm system design

    International Nuclear Information System (INIS)

    Greenfield, B.

    2009-01-01

    An instructional manual was created to guide criticality safety engineers through the technical aspects of designing a criticality alarm system (CAS) for Dept. of Energy (DOE) hazard class 1 and 2 facilities. The manual was structured such that it can be used by engineers designing completely new systems and by those who are working with existing facilities. Major design tasks are thoroughly analyzed to provide concise direction for how to complete the analysis. Regulatory and technical performance requirements were both addressed. (authors)

  10. 21 CFR 870.1100 - Blood pressure alarm.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Blood pressure alarm. 870.1100 Section 870.1100...) MEDICAL DEVICES CARDIOVASCULAR DEVICES Cardiovascular Diagnostic Devices § 870.1100 Blood pressure alarm. (a) Identification. A blood pressure alarm is a device that accepts the signal from a blood pressure...

  11. 46 CFR 78.47-75 - Ventilation alarm failure.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 3 2010-10-01 2010-10-01 false Ventilation alarm failure. 78.47-75 Section 78.47-75... Fire and Emergency Equipment, Etc. § 78.47-75 Ventilation alarm failure. (a) The alarm required by § 72... FAILURE IN VEHICULAR SPACE.” (b) [Reserved] [CGFR 66-33, 31 FR 15284, Dec. 6, 1966] ...

  12. The sound of danger: threat sensitivity to predator vocalizations, alarm calls, and novelty in gulls.

    Directory of Open Access Journals (Sweden)

    Sarah A MacLean

    Full Text Available The threat sensitivity hypothesis predicts that organisms will evaluate the relative danger of and respond differentially to varying degrees of predation threat. Doing so allows potential prey to balance the costs and benefits of anti-predator behaviors. Threat sensitivity has undergone limited testing in the auditory modality, and the relative threat level of auditory cues from different sources is difficult to infer across populations when variables such as background risk and experience are not properly controlled. We experimentally exposed a single population of two sympatric gull species to auditory stimuli representing a range of potential threats in order to compare the relative threat of heterospecific alarm calls, conspecific alarms calls, predator vocalizations, and novel auditory cues. Gulls were able to discriminate among a diverse set of threat indicators and respond in a graded manner commensurate with the level of threat. Vocalizations of two potential predators, the human voice and bald eagle call, differed in their threat level compared to each other and to alarm calls. Conspecific alarm calls were more threatening than heterospecfic alarm calls to the larger great black-backed gull, but the smaller herring gull weighed both equally. A novel cue elicited a response intermediate between known threats and a known non-threat in herring gulls, but not great black-backed gulls. Our results show that the relative threat level of auditory cues from different sources is highly species-dependent, and that caution should be exercised when comparing graded and threshold threat sensitive responses.

  13. Recommendations to alarm systems and lessons learned on alarm system implementation

    International Nuclear Information System (INIS)

    Soerenssen, Aimar; Veland, Oeystein; Farbrot, Jan Erik; Kaarstad, Magnhild; Seim, Lars Aage; Foerdestroemmen, Nils; Bye, Andreas

    2001-11-01

    Alarm systems have been of major concern within complex industrial processes for many years. Within the nuclear community, the TMI accident in 1979 was the first really serious event that showed also the importance of the man-machine aspects of the systems in general, and the alarm system in particular. The OECD Halden Reactor Project has been working with alarm systems since 1974. This report is an attempt to gather some of the knowledge that has been accumulated during the years in Halden, both in research and also in bilateral projects. Bilateral projects within this field have provided a practical basis of knowledge.A major part of this report consists of a set of recommendations, which reflect HRP's current understanding of how an alarm system should work. There are also recommendations on design methods. But also other issues are included, as system development and implementation experience, and experimental knowledge on the performance of alarm systems. Some open issues are also discussed. (Author). 54 refs., 15 figs

  14. Nitric Acid Revamp and Upgrading of the Alarm & Protection Safety System at Petrokemija, Croatia

    Directory of Open Access Journals (Sweden)

    Hoško, I.

    2012-04-01

    Full Text Available Every industrial production, particularly chemical processing, demands special attention in conducting the technological process with regard to the security requirements. For this reason, production processes should be continuously monitored by means of control and alarm safety instrumented systems. In the production of nitric acid at Petrokemija d. d., the original alarm safety system was designed as a combination of an electrical relay safety system and transistorized alarm module system. In order to increase safety requirements and modernize the technological process of nitric acid production, revamping and upgrading of the existing alarm safety system was initiated with a new microprocessor system. The newly derived alarm safety system, Simatic PCS 7, links the function of "classically" distributed control (DCS and logical systems in a common hardware and software platform with integrated engineering tools and operator interface to meet the minimum safety standards with safety integrity level 2 (SIL2 up to level 3 (SIL3, according to IEC 61508 and IEC 61511. This professional paper demonstrates the methodology of upgrading the logic of the alarm safety system in the production of nitric acid in the form of a logical diagram, which was the basis for a further step in its design and construction. Based on the mentioned logical diagram and defined security requirements, the project was implemented in three phases: analysis and testing, installation of the safety equipment and system, and commissioning. Developed also was a verification system of all safety conditions, which could be applied to other facilities for production of nitric acid. With the revamped and upgraded interlock alarm safety system, a new and improved safety boundary in the production of nitric acid was set, which created the foundation for further improvement of the production process in terms of improved analysis.

  15. Stream Habitat Reach Summary - NCWAP [ds158

    Data.gov (United States)

    California Natural Resource Agency — The Stream Habitat - NCWAP - Reach Summary [ds158] shapefile contains in-stream habitat survey data summarized to the stream reach level. It is a derivative of the...

  16. A Study on Performance Requirements for Advanced Alarm System

    International Nuclear Information System (INIS)

    Seong, Duk Hyun; Jeong, Jae Hoon; Sim, Young Rok; Ko, Jong Hyun; Kim, Jung Seon; Jang, Gwi Sook; Park, Geun Ok

    2005-01-01

    A design goals of advanced alarm system is providing advanced alarm information to operator in main control room. To achive this, we applied computer based system to Alarm System. Because, It should apply data management and advanced alarm processing(ie. Data Base Mangegment System and S/W module for alarm processing). These are not impossible in analog based alarm system. And, preexitance research examples are made on digital computer. We have digital systems for test of advanced alarm system table and have tested and studied using by test equipment in the view point of the system performance, stability and security. In this paper, we discribed about general software architecture of preexitance research examples. Also, CPU performance and requirements of system software that served to accommodate it, stability and security

  17. Applying AI techniques to improve alarm display effectiveness

    International Nuclear Information System (INIS)

    Gross, J.M.; Birrer, S.A.; Crosberg, D.R.

    1987-01-01

    The Alarm Filtering System (AFS) addresses the problem of information overload in a control room during abnormal operations. Since operators can miss vital information during these periods, systems which emphasize important messages are beneficial. AFS uses the artificial intelligence (AI) technique of object-oriented programming to filter and dynamically prioritize alarm messages. When an alarm's status changes, AFS determines the relative importance of that change according to the current process state. AFS bases that relative importance on relationships the newly changed alarm has with other activated alarms. Evaluations of a alarm importance take place without regard to the activation sequence of alarm signals. The United States Department of Energy has applied for a patent on the approach used in this software. The approach was originally developed by EG and G Idaho for a nuclear reactor control room

  18. An experimental evaluation of alarm processing and display characteristics

    International Nuclear Information System (INIS)

    O'Hara, J.; Brown, W.; Hallbert, B.; Skraaning, G.Jr.; Persensky, J.; Wachtel, J.

    1998-01-01

    This paper describes a research program sponsored by the U.S. Nuclear Regulatory Commission to address the human factors engineering (HFE) aspects of nuclear power plant alarm systems. The overall objective of the program is to develop HFE review guidance for advanced alarm systems. As part of this program, guidance has been developed based on a broad base of technical and research literature. In the course of guidance development, aspects of alarm system design for which the technical basis was insufficient to support complete guidance development were identified. The primary purpose of the research reported in this paper was to evaluate the effects of three of these alarm system design characteristics on operator performance in order to contribute to the understanding of potential safety issues and to provide data to support the development of design review guidance in these areas. Three alarm system design characteristics studied were (1) alarm processing (degree of alarm reduction), (2) alarm availability (dynamic prioritization and suppression), and (3) alarm display (a dedicated tile format, a mixed tile and message list format, and a format in which alarm information is integrated into the process displays). A secondary purpose was to provide confirmatory evidence of selected alarm system guidance developed in an earlier phase of the project. The alarm characteristics were combined into eight separate experimental conditions. Six, two-person crews of professional nuclear power plant operators participated in the study. Following training, each crew completed 16 test trials which consisted of two trials in each of the eight experimental conditions (one with a low-complexity scenario and one with a high-complexity scenario). Measures of process performance. operator task performance, situation awareness, and workload were obtained. In addition. operator opinions and evaluations of the alarm processing and display conditions were collected. Numerous strengths

  19. Teratology testing under REACH.

    Science.gov (United States)

    Barton, Steve

    2013-01-01

    REACH guidelines may require teratology testing for new and existing chemicals. This chapter discusses procedures to assess the need for teratology testing and the conduct and interpretation of teratology tests where required.

  20. Global reach and engagement

    Science.gov (United States)

    2016-09-01

    Popular culture reflects both the interests of and the issues affecting the general public. As concerns regarding climate change and its impacts grow, is it permeating into popular culture and reaching that global audience?

  1. Security alarm communication and display systems development

    International Nuclear Information System (INIS)

    Waddoups, I.G.

    1990-01-01

    Sandia National Laboratories (SNL) has, as lead Department of Energy (DOE) physical security laboratory, developed a variety of alarm communication and display systems for DOE and Department of Defense (DOD) facilities. This paper briefly describes some of the systems developed and concludes with a discussion of technology relevant to those currently designing, developing, implementing, or procuring such a system. Development activities and the rapid evolution of computers over the last decade have resulted in a broad variety of capabilities to support most security system communication and display needs. The major task in selecting a system is becoming familiar with these capabilities and finding the best match to a specific need

  2. Alarm annunciation in a graphical environment

    International Nuclear Information System (INIS)

    Adams, D.G.

    1994-01-01

    Well-designed graphical user interfaces, such as Microsoft reg-sign Windows trademark or UNIX trademark--based X-Windows reg-sign, provide a capability for enhanced display of security alarm information. Conversely, a poorly designed interface can quickly overwhelm an operator. This paper describes types of graphical information that can be displayed and offers guidance on how to best display that information. Limits are proposed for the complexity of the user interface, and guidelines are suggested for the display of maps and sensors

  3. Monitoring techniques and alarm procedures for CMS services and sites in WLCG

    Energy Technology Data Exchange (ETDEWEB)

    Molina-Perez, J. [UC, San Diego; Bonacorsi, D. [Bologna U.; Gutsche, O. [Fermilab; Sciaba, A. [CERN; Flix, J. [Madrid, CIEMAT; Kreuzer, P. [CERN; Fajardo, E. [Andes U., Bogota; Boccali, T. [INFN, Pisa; Klute, M. [MIT; Gomes, D. [Rio de Janeiro State U.; Kaselis, R. [Vilnius U.; Du, R. [Beijing, Inst. High Energy Phys.; Magini, N. [CERN; Butenas, I. [Vilnius U.; Wang, W. [Beijing, Inst. High Energy Phys.

    2012-01-01

    The CMS offline computing system is composed of roughly 80 sites (including most experienced T3s) and a number of central services to distribute, process and analyze data worldwide. A high level of stability and reliability is required from the underlying infrastructure and services, partially covered by local or automated monitoring and alarming systems such as Lemon and SLS, the former collects metrics from sensors installed on computing nodes and triggers alarms when values are out of range, the latter measures the quality of service and warns managers when service is affected. CMS has established computing shift procedures with personnel operating worldwide from remote Computing Centers, under the supervision of the Computing Run Coordinator at CERN. This dedicated 24/7 computing shift personnel is contributing to detect and react timely on any unexpected error and hence ensure that CMS workflows are carried out efficiently and in a sustained manner. Synergy among all the involved actors is exploited to ensure the 24/7 monitoring, alarming and troubleshooting of the CMS computing sites and services. We review the deployment of the monitoring and alarming procedures, and report on the experience gained throughout the first two years of LHC operation. We describe the efficiency of the communication tools employed, the coherent monitoring framework, the proactive alarming systems and the proficient troubleshooting procedures that helped the CMS Computing facilities and infrastructure to operate at high reliability levels.

  4. Monitoring techniques and alarm procedures for CMS Services and Sites in WLCG

    International Nuclear Information System (INIS)

    Molina-Perez, J; Sciabà, A; Magini, N; Bonacorsi, D; Gutsche, O; Flix, J; Kreuzer, P; Fajardo, E; Boccali, T; Klute, M; Gomes, D; Kaselis, R; Butenas, I; Du, R; Wang, W

    2012-01-01

    The CMS offline computing system is composed of roughly 80 sites (including most experienced T3s) and a number of central services to distribute, process and analyze data worldwide. A high level of stability and reliability is required from the underlying infrastructure and services, partially covered by local or automated monitoring and alarming systems such as Lemon and SLS; the former collects metrics from sensors installed on computing nodes and triggers alarms when values are out of range, the latter measures the quality of service and warns managers when service is affected. CMS has established computing shift procedures with personnel operating worldwide from remote Computing Centers, under the supervision of the Computing Run Coordinator at CERN. This dedicated 24/7 computing shift personnel is contributing to detect and react timely on any unexpected error and hence ensure that CMS workflows are carried out efficiently and in a sustained manner. Synergy among all the involved actors is exploited to ensure the 24/7 monitoring, alarming and troubleshooting of the CMS computing sites and services. We review the deployment of the monitoring and alarming procedures, and report on the experience gained throughout the first two years of LHC operation. We describe the efficiency of the communication tools employed, the coherent monitoring framework, the proactive alarming systems and the proficient troubleshooting procedures that helped the CMS Computing facilities and infrastructure to operate at high reliability levels.

  5. Interspecific semantic alarm call recognition in the solitary Sahamalaza sportive lemur, Lepilemur sahamalazensis.

    Directory of Open Access Journals (Sweden)

    Melanie Seiler

    Full Text Available As alarm calls indicate the presence of predators, the correct interpretation of alarm calls, including those of other species, is essential for predator avoidance. Conversely, communication calls of other species might indicate the perceived absence of a predator and hence allow a reduction in vigilance. This "eavesdropping" was demonstrated in birds and mammals, including lemur species. Interspecific communication between taxonomic groups has so far been reported in some reptiles and mammals, including three primate species. So far, neither semantic nor interspecific communication has been tested in a solitary and nocturnal lemur species. The aim of this study was to investigate if the nocturnal and solitary Sahamalaza sportive lemur, Lepilemur sahamalazensis, is able to access semantic information of sympatric species. During the day, this species faces the risk of falling prey to aerial and terrestrial predators and therefore shows high levels of vigilance. We presented alarm calls of the crested coua, the Madagascar magpie-robin and aerial, terrestrial and agitation alarm calls of the blue-eyed black lemur to 19 individual Sahamalaza sportive lemurs resting in tree holes. Songs of both bird species' and contact calls of the blue-eyed black lemur were used as a control. After alarm calls of crested coua, Madagascar magpie-robin and aerial alarm of the blue-eyed black lemur, the lemurs scanned up and their vigilance increased significantly. After presentation of terrestrial alarm and agitation calls of the blue-eyed black lemur, the animals did not show significant changes in scanning direction or in the duration of vigilance. Sportive lemur vigilance decreased after playbacks of songs of the bird species and contact calls of blue-eyed black lemurs. Our results indicate that the Sahamalaza sportive lemur is capable of using information on predator presence as well as predator type of different sympatric species, using their referential

  6. Does Digital Video Advertising Increase Population-Level Reach of Multimedia Campaigns? Evidence From the 2013 Tips From Former Smokers Campaign

    OpenAIRE

    Davis, Kevin C; Shafer, Paul R; Rodes, Robert; Kim, Annice; Hansen, Heather; Patel, Deesha; Coln, Caryn; Beistle, Diane

    2016-01-01

    Background Federal and state public health agencies in the United States are increasingly using digital advertising and social media to promote messages from broader multimedia campaigns. However, little evidence exists on population-level campaign awareness and relative cost efficiencies of digital advertising in the context of a comprehensive public health education campaign. Objective Our objective was to compare the impact of increased doses of digital video and television advertising fro...

  7. Does Digital Video Advertising Increase Population-Level Reach of Multimedia Campaigns? Evidence From the 2013 Tips From Former Smokers Campaign.

    Science.gov (United States)

    Davis, Kevin C; Shafer, Paul R; Rodes, Robert; Kim, Annice; Hansen, Heather; Patel, Deesha; Coln, Caryn; Beistle, Diane

    2016-09-14

    Federal and state public health agencies in the United States are increasingly using digital advertising and social media to promote messages from broader multimedia campaigns. However, little evidence exists on population-level campaign awareness and relative cost efficiencies of digital advertising in the context of a comprehensive public health education campaign. Our objective was to compare the impact of increased doses of digital video and television advertising from the 2013 Tips From Former Smokers (Tips) campaign on overall campaign awareness at the population level. We also compared the relative cost efficiencies across these media platforms. We used data from a large national online survey of approximately 15,000 US smokers conducted in 2013 immediately after the conclusion of the 2013 Tips campaign. These data were used to compare the effects of variation in media dose of digital video and television advertising on population-level awareness of the Tips campaign. We implemented higher doses of digital video among selected media markets and randomly selected other markets to receive similar higher doses of television ads. Multivariate logistic regressions estimated the odds of overall campaign awareness via digital or television format as a function of higher-dose media in each market area. All statistical tests used the .05 threshold for statistical significance and the .10 level for marginal nonsignificance. We used adjusted advertising costs for the additional doses of digital and television advertising to compare the cost efficiencies of digital and television advertising on the basis of costs per percentage point of population awareness generated. Higher-dose digital video advertising was associated with 94% increased odds of awareness of any ad online relative to standard-dose markets (Pdigital advertising was associated with a marginally nonsignificant increase (46%) in overall campaign awareness regardless of media format (P=.09). Higher

  8. Does Digital Video Advertising Increase Population-Level Reach of Multimedia Campaigns? Evidence From the 2013 Tips From Former Smokers Campaign

    Science.gov (United States)

    Shafer, Paul R; Rodes, Robert; Kim, Annice; Hansen, Heather; Patel, Deesha; Coln, Caryn; Beistle, Diane

    2016-01-01

    Background Federal and state public health agencies in the United States are increasingly using digital advertising and social media to promote messages from broader multimedia campaigns. However, little evidence exists on population-level campaign awareness and relative cost efficiencies of digital advertising in the context of a comprehensive public health education campaign. Objective Our objective was to compare the impact of increased doses of digital video and television advertising from the 2013 Tips From Former Smokers (Tips) campaign on overall campaign awareness at the population level. We also compared the relative cost efficiencies across these media platforms. Methods We used data from a large national online survey of approximately 15,000 US smokers conducted in 2013 immediately after the conclusion of the 2013 Tips campaign. These data were used to compare the effects of variation in media dose of digital video and television advertising on population-level awareness of the Tips campaign. We implemented higher doses of digital video among selected media markets and randomly selected other markets to receive similar higher doses of television ads. Multivariate logistic regressions estimated the odds of overall campaign awareness via digital or television format as a function of higher-dose media in each market area. All statistical tests used the .05 threshold for statistical significance and the .10 level for marginal nonsignificance. We used adjusted advertising costs for the additional doses of digital and television advertising to compare the cost efficiencies of digital and television advertising on the basis of costs per percentage point of population awareness generated. Results Higher-dose digital video advertising was associated with 94% increased odds of awareness of any ad online relative to standard-dose markets (Pdigital advertising was associated with a marginally nonsignificant increase (46%) in overall campaign awareness regardless of

  9. A Design of Alarm System in a Research Reactor Facility

    International Nuclear Information System (INIS)

    Park, Jaekwan; Jang, Gwisook; Seo, Sangmun; Suh, Yongsuk

    2013-01-01

    The digital alarm system has become an indispensable design to process a large amount of alarms of power plants. Korean research reactor operated for decades maintains a hybrid alarm system with both an analog annunciator and a digital alarm display. In this design, several alarms are indicated on an analog panel and digital display, respectively, and it requires more attention and effort of the operators. As proven in power plants, a centralized alarm system design is necessary for a new research reactor. However, the number of alarms and operators in a research reactor is significantly lesser than power plants. Thus, simplification should be considered as an important factor for the operation efficiency. This paper introduces a simplified alarm system. As advances in information technology, fully digitalized alarm systems have been applied to power plants. In a new research reactor, it will be more useful than an analog or hybrid configuration installed in research reactors decades ago. However, the simplification feature should be considered as an important factor because the number of alarms and number of operators in a research reactor is significantly lesser than in power plants

  10. JOYO operation support system 'JOYCAT' based on intelligent alarm handling

    International Nuclear Information System (INIS)

    Tamaoki, Tetsuo; Yamamoto, Hiroki; Sato, Masuo; Yoshida, Megumu; Kaneko, Tomoko; Terunuma, Seiichi; Takatsuto, Hiroshi; Morimoto, Makoto.

    1992-01-01

    An operation support system for the experimental fast reactor 'JOYO' was developed based on an intelligent alarm-handling. A specific feature of this system, called JOYCAT (JOYO Consulting and Analyzing Tool), is in its sequential processing structure that a uniform treatment by using design knowledge base is firstly applied for all activated alarms, and an exceptional treatment by using heuristic knowledge base is then applied only for the former results. This enables us to achieve real-time and flexible alarm-handling. The first alarm-handling determines the candidates of causal alarms, important alarms with which the operator should firstly cope, through identifying the cause-consequence relations among alarms based on the design knowledge base in which importance and activating conditions are described for each of 640 alarms in a frame format. The second alarm-handling makes the final judgement with the candidates by using the heuristic knowledge base described as production rules. Then, operation manuals concerning the most important alarms are displayed to operators. JOYCAT has been in commission since September of 1990, after a wide scope of validation tests by using an on-site full-scope training simulator. (author)

  11. Alarm management in gas pipeline plant: a case study

    Energy Technology Data Exchange (ETDEWEB)

    Araujo, Juliano; Lima, Marcelo; Leitao, Gustavo; Guedes, Luiz Affonso [Universidade Federal do Rio Grande do Norte (UFRN), Natal, RN (Brazil); Branco, Nicolau; Coelho, Robson; Elias, Gustavo Passos; Nunes, Marcelo [Transportadora Brasileira Gasoduto Bolivia-Brasil (TBG), Rio de Janeiro, RJ (Brazil)

    2009-07-01

    In order to improve the requirements of industrial processes, many decision support systems have been introduced in recent years. In this context, the alarm management systems have great relevance. On the other hand, the informatics revolution allowed a great increase of information concerning the operation of the industrial processes. Currently, process operators handle an excessive number of about 1.500 alarms per day. Thus, this overdose of information implies in the discredit of alarms. Then, in order to improve the operation activities of industrial processes, it is mandatory to incorporate procedures to evaluate and rationalize alarms. Since the EMMUA191 Standard is the reference guide to alarm management, but it does not specify how to execute an alarm management procedure, in this paper, a systematic procedure to evaluate alarms configurations in industrial processes is proposed. This procedure is in line with EMMUA191 and is composed by the following steps: to use statistics analyses to identify problematic alarms, such as occurrence, intermittency, correlation, and flooding calculation; to indicate problematic alarm group; and to propose a set of actions to be implemented. To validate our proposal, we present a case study in a gas pipeline plant using the BR-AlarmExpert software. (author)

  12. Framework for combining REACH and national regulations to obtain equal protection levels of human health and the environment in different countries - Comparative study of Denmark and Korea

    DEFF Research Database (Denmark)

    Lee, Jihyun; Pedersen, Anders Branth; Thomsen, Marianne

    2013-01-01

    The aim of this paper is to present a conceptual framework for a systems approach to protect the environment and human health by taking into account differences in the cumulative risks of total human exposure in a territorial context. To this end the measures that are available and that can...... compound, the territorial differences in background exposure to endocrine and neurological interfering stressors were modelled. It is concluded that the different territorial soil and air lead pollution levels contribute differently to the total childhood lead exposure in the two countries. As such...

  13. General methods for alarm reduction; Larmsanering med generella metoder

    Energy Technology Data Exchange (ETDEWEB)

    Ahnlund, Jonas; Bergquist, Tord; Raaberg, Martin [Lund Univ. (Sweden). Dept. of Information Technology

    2003-10-01

    The information in the control rooms has increased due to the technological advances in process control. Large industries produce large data quantities, where some information is unnecessary or even incorrect. The operator needs support from an advanced and well-adjusted alarm system to be able to separate a real event from a minor disturbance. The alarms must be of assistance and not a nuisance. An enhanced alarm situation qualifies an increased efficiency with fewer production disturbances and an improved safety. Yet, it is still unusual that actions are taken to improve the situation. An alarm cleanup with general methods can shortly be described as taking advantage of the control systems built-in functions, the possibility to modify or create function blocks and fine-tune the settings in the alarm system. In this project, we make use of an intelligent software, Alarm Cleanup Toolbox, that simulate different signal processing methods and tries to find improved settings on all the signals in the process. This is a fast and cost-efficient way to improve the overall alarm situation, and lays a foundation for more advanced alarm systems. An alarm cleanup has been carried out at Flintraennan district heating plant in Malmoe, where various signal processing methods has been implemented in a parallel alarm system. This made it possible to compare the two systems under the same conditions. The result is very promising, and shows that a lot of improvements can be achieved with very little effort. An analysis of the alarm system at Vattenreningen (the water purification process) at Heleneholmsverket in Malmoe has been carried out. Alarm Cleanup Toolbox has, besides suggesting improved settings, also found logical errors in the alarm system. Here, no implementation was carried out and therefore the results are analytical, but they validate the efficiency of the general methods. The project has shown that an alarm cleanup with general methods is cost-efficient, and that the

  14. Urban density and the metabolic reach of metropolitan areas: A panel analysis of per capita transportation emissions at the county-level.

    Science.gov (United States)

    Ergas, Christina; Clement, Matthew; McGee, Julius

    2016-07-01

    We engage a tension in the urban environment literature that positions cities as both drivers of environmental destruction and loci of environmental protection. We argue that the traditional binary view of cities as either harmful or beneficial is too simplistic; we advance a more nuanced understanding of cities to study their internal and external metabolic effects in terms of carbon emissions from on-road transportation at the county-level across the continental United States between 2002 and 2007. First, utilizing satellite imagery from the National Land Cover Database, we create a novel measure of population density by quantifying the number of people per square mile of impervious surface area. Second, we develop a measure of metropolitan adjacency from the rural classifications datasets published by the USDA. In spatial regression models, we find that while higher density reduces emissions, counties that are geographically isolated from metropolitan areas actually have lower per capita emissions, all else equal. We elaborate on the conceptual, methodological, and practical implications of our study in the conclusion. Copyright © 2016 Elsevier Inc. All rights reserved.

  15. Classification of alarm processing techniques and human performance issues

    International Nuclear Information System (INIS)

    Kim, I.S.; O'Hara, J.M.

    1993-01-01

    Human factors reviews indicate that conventional alarm systems based on the one sensor, one alarm approach, have many human engineering deficiencies, a paramount example being too many alarms during major disturbances. As an effort to resolve these deficiencies, various alarm processing systems have been developed using different techniques. To ensure their contribution to operational safety, the impacts of those systems on operating crew performance should be carefully evaluated. This paper briefly reviews some of the human factors research issues associated with alarm processing techniques and then discusses a framework with which to classify the techniques. The dimensions of this framework can be used to explore the effects of alarm processing systems on human performance

  16. Experimental evaluation of human-system interaction on alarm design

    International Nuclear Information System (INIS)

    Huang, F.-H.; Lee, Y.-L.; Hwang, S.-L.; Yenn, T.-C.; Yu, Y.-C.; Hsu, C.-C.; Huang, H.-W.

    2007-01-01

    This study evaluates the practicability of automatic reset alarm system in Fourth Nuclear Power Plant (FNPP) of Taiwan. The features of auto-reset alarm system include dynamic prioritization of all alarm signals and fast system reset. Two experiments were conducted to evaluate the effect of automatic/manual reset on operation time, situational awareness (SA), task load index (TLX), and subjective ratings. All participants, including Experts and Novices, took part in the experiment on the alarm system simulator with Load Rejection procedure. The experimental results imply that the auto-reset alarm system may be applied in an advanced control room under Load Rejection procedure, because all participants' operation time were reduced as well as Novice's SA were raised up. Nevertheless, to ensure operating safety in FNPP, the effects of the auto-reset alarm system in other procedures/special situations still need to be tested in the near future

  17. Framework for analyzing safeguards alarms and response decisions

    International Nuclear Information System (INIS)

    Al-Ayat, R.A.; Judd, B.R.; McCord, R.K.

    1982-01-01

    This paper describes a quantitative approach to help evaluate and respond to safeguards alarms. These alrms may be generated internally by a facility's safeguards systems or externally by individuals claiming to possess stolen Special Nuclear Material (SNM). This approach can be used to identify the most likely cause of an alarm - theft, hoax, or error - and to evaluate alternative responses to alarms. Possible responses include conducting investigations, initiating measures to recover stolen SNM, and replying to external threats. Based on the results of each alarm investigation step, the evaluation revises the likelihoods of possible causes of an alarm, and uses this information to determine the optimal sequence of further responses. The choice of an optimal sequence of responses takes into consideration the costs and benefits of successful thefts or hoaxes. These results provide an analytical basis for setting priorities and developing contingency plans for responding to safeguards alarms

  18. Classification of alarm processing techniques and human performance issues

    Energy Technology Data Exchange (ETDEWEB)

    Kim, I.S.; O' Hara, J.M.

    1993-01-01

    Human factors reviews indicate that conventional alarm systems based on the one sensor, one alarm approach, have many human engineering deficiencies, a paramount example being too many alarms during major disturbances. As an effort to resolve these deficiencies, various alarm processing systems have been developed using different techniques. To ensure their contribution to operational safety, the impacts of those systems on operating crew performance should be carefully evaluated. This paper briefly reviews some of the human factors research issues associated with alarm processing techniques and then discusses a framework with which to classify the techniques. The dimensions of this framework can be used to explore the effects of alarm processing systems on human performance.

  19. Classification of alarm processing techniques and human performance issues

    Energy Technology Data Exchange (ETDEWEB)

    Kim, I.S.; O`Hara, J.M.

    1993-05-01

    Human factors reviews indicate that conventional alarm systems based on the one sensor, one alarm approach, have many human engineering deficiencies, a paramount example being too many alarms during major disturbances. As an effort to resolve these deficiencies, various alarm processing systems have been developed using different techniques. To ensure their contribution to operational safety, the impacts of those systems on operating crew performance should be carefully evaluated. This paper briefly reviews some of the human factors research issues associated with alarm processing techniques and then discusses a framework with which to classify the techniques. The dimensions of this framework can be used to explore the effects of alarm processing systems on human performance.

  20. Resolution of alarms for loss of bulk nuclear material

    International Nuclear Information System (INIS)

    Eggers, R.F.; Davenport, L.C.

    1982-01-01

    Under methods of material accountability considered in the NRC's Reform Amendment (Federal Register, 46(175):45144 to 45151 dated September 10, 1981) prompt detection of losses and resolution of alarms play a central role in the day-to-day activities of the Material Control and Accounting (MC and A) System. This paper will discuss the two basic pathways of alarm resolution, namely, verification of the magnitude of the loss indicated by the initial alarm, and detection of deliberate or accidental accounting discrepancies. Progress along these pathways leads to a consensus that either (1) a loss occurred, (2) the original alarm was caused by MC and A error, or (3) the cause of the original alarm is uncertain. Three phases of response will be outlined and an example of response to alarms will be given for a mixed oxide powder processing control unit

  1. CRISIS UNDER THE RADAR: ILLICIT AMPHETAMINE USE IS REACHING EPIDEMIC PROPORTIONS AND CONTRIBUTING TO RESOURCE OVER-UTILIZATION AT A LEVEL 1 TRAUMA CENTER.

    Science.gov (United States)

    Gemma, Vincent A; Chapple, Kristina A; Goslar, Pamela W; Israr, Sharjeel; Petersen, Scott R; Weinberg, Jordan A

    2018-05-21

    Epidemiological LEVEL OF EVIDENCE: III.

  2. Expert System Constant False Alarm Rate (CFAR) Processor

    National Research Council Canada - National Science Library

    Wicks, Michael C

    2006-01-01

    An artificial intelligence system improves radar signal processor performance by increasing target probability of detection and reducing probability of false alarm in a severe radar clutter environment...

  3. Operator Performance Comparison of two VDT-based Alarm Systems

    International Nuclear Information System (INIS)

    Lee, Hyun-Chul; Oh, In-Suk; Sim, Bong-Shick; Koo, In-Soo; Kim, Jeong-Taek; Lee, Ki-Young; Park, Jong-Kyun

    1998-01-01

    This study is carried out to investigate performance differences between two alarm presentation methods from the viewpoint of human factors and to provide items to be improved. One of the alarm display methods considered in this study displays alarm lists on VDT combined with hardwired alarm panels. The other method displays alarms on plant mimic diagrams of VDT. This alarm display method has other features for operator aid with which operator can get detailed information on the activated alarm in the mimic diagrams, and the capability for alarm processing such as alarm reduction and prioritization. To compare the two display methods, a human factor experiment was performed with a plant simulator in the ITF (Integrated Test Facility) that plant operators run for 4 event scenarios. During the experiment, physiological measurements, system and operator action log, and audio/video recordings were collected. Operators subjective opinion was collected as well after the experiment. Time, error rate and situation awareness were major human factor criteria used for the comparison during the analysis stage of the experiment. No statistical significance was found in the results of our statistical comparison analysis. Several findings were identified, however, through the analysis of subjective opinions. (authors)

  4. Contribution of computerization to alarm processing: A French safety view

    Energy Technology Data Exchange (ETDEWEB)

    Cette, W [CEA Centre d` Etudes de Fontenay-aux-Roses, 92 (France). Inst. de Protection et de Surete Nucleaire

    1997-09-01

    Following the TMI accident and according to the requirement of the French safety authority, very important studies were performed by the French utility, Electricite de France (EDF), and assessed by the Institute for Nuclear Safety and Protection (IPSN) on reactor operation in conventional control rooms, particularly on alarm processing. These studies dealt with the man-machine interface, as well as design and exploitation requirements, presentation and management of alarm signals, and associated operating documents. The conclusions of these studies have led to improvements in French conventional control rooms. The current state of these control rooms and links between alarm sets and operating documents will be shortly presented in the first part of the paper. More recently, the computerized means implemented in the PWR 1400 MWe control rooms (N4) profoundly modified reactor operation. In particular, major advances concern alarm processing in comparison with conventional control rooms. The N4 plants provide a more rigorous approach in processing and presentation of alarms than in the past. Indeed, EDF wanted to have less alarms switched on during plant upsets and to make them more characteristic of a specific situation of the process. For example, computerization makes it easier to validate or inhibit alarms according to the situation, to allow the operator to manage alarm presentation and to propose on-line alarm sheets to the operator etc. This approach in comparison with conventional control rooms, and the IPSN assessment will be presented in the second part of this paper. (author).

  5. AI-based alarm processing for a nuclear power plant

    International Nuclear Information System (INIS)

    Na, N.J.; Kim, I.S.; Hwang, I.K.; Lee, D.Y.; Ham, C.S.

    1996-01-01

    A real-time expert system is implemented using artificial intelligence and object-oriented technology for alarm processing and presentation in a nuclear power plant. The knowledge base is constructed based on some schemes to process and display alarms to the plant operators. The activated alarms are dynamically prioritized by the reasoning rules, and then, presented on the process mimic overview and by some other means. To demonstrate the proposed system, the alarm processing and presentation is carried out in a simulated environment of the TMI-2 accident

  6. Contribution of computerization to alarm processing: A French safety view

    International Nuclear Information System (INIS)

    Cette, W.

    1997-01-01

    Following the TMI accident and according to the requirement of the French safety authority, very important studies were performed by the French utility, Electricite de France (EDF), and assessed by the Institute for Nuclear Safety and Protection (IPSN) on reactor operation in conventional control rooms, particularly on alarm processing. These studies dealt with the man-machine interface, as well as design and exploitation requirements, presentation and management of alarm signals, and associated operating documents. The conclusions of these studies have led to improvements in French conventional control rooms. The current state of these control rooms and links between alarm sets and operating documents will be shortly presented in the first part of the paper. More recently, the computerized means implemented in the PWR 1400 MWe control rooms (N4) profoundly modified reactor operation. In particular, major advances concern alarm processing in comparison with conventional control rooms. The N4 plants provide a more rigorous approach in processing and presentation of alarms than in the past. Indeed, EDF wanted to have less alarms switched on during plant upsets and to make them more characteristic of a specific situation of the process. For example, computerization makes it easier to validate or inhibit alarms according to the situation, to allow the operator to manage alarm presentation and to propose on-line alarm sheets to the operator etc. This approach in comparison with conventional control rooms, and the IPSN assessment will be presented in the second part of this paper. (author)

  7. Testing alarm resolution procedures in a fuel fabrication facility

    International Nuclear Information System (INIS)

    Smith, B.W.; Razvi, J.

    1984-07-01

    Process monitoring data can be used for generating material loss estimates. The intent of using process control data is to enhance nuclear material control and accounting for the timely detection and resolution of discrepancies. The purpose of an alarm resolution system is to distinguish between system errors and an actual loss of nuclear material. A study has been performed to develop and test a site-specific set of alarm resolution procedures. The results of the study are described and include the frequency of alarms, the causes of alarms, the type of resolution, and the modeling of loss estimates. 3 references, 2 figures, 2 tables

  8. Kidkit guides children into alarming atmospheres

    DEFF Research Database (Denmark)

    Højlund, Marie; Kinch, Sofie

    2013-01-01

    This paper presents the concept of Embodied Habituation as an architectural approach to designing contextualized technologies. It does so by identifying Middle Ground Experiences acknowledging how spaces are inhabited with ambiguous qualities that affect people emotionally. The research is based...... on the development and evaluation of Kidkit, which is interactive furniture designed for young children who are going to visit a hospitalized relative with fatal injuries for the first time. Kidkit empowers the child to engage and be present by shaping Middle Ground Experiences in the hospital ward environment...... that is full of intimidating medical equipment and alarms. The evaluation results indicate collective rewards gained when children succeed in Embodied Habituation. Finally, the paper discusses how Middle Ground Experiences inevitably establish grounds for how we design for spatial experiences within...

  9. Aplikasi Sensor Cahaya Untuk Alarm Anti Pencuri

    Directory of Open Access Journals (Sweden)

    Asita Shoman Muzaki

    2011-11-01

    Full Text Available Kasus pencurian di rumah kosong yang ditinggal pergi oleh pemiliknya belakangan ini marak terjadi. Berangkat dari pemikiran ini penulis mencoba merancang alarm yang dapat mendeteksi pergerakan seseorang saat rumah dalam kondisi kosong, ditinggalkan oleh pemiliknya. Alat ini mempunyai prinsip kerja yaitu mendeteksi bayangan seseorang yang melewati titik tertentu. Perancangan dan pembuatan perangkat ini menggunakan sensor cahaya berupa LASER dan LDR yang dirangkai dengan transistor sebagai saklar otomatis serta LED dan telepon rumah untuk melakukan panggilan kepada nomor telepon pemilik rumah. Komponen yang dipakai dalam pembuatan perangkat ini antara lain IC LM7805, LASER pointer, resistor, transistor BC108, LED, relay dan telepon rumah. Perancangan dan pembuatan alat menggunakan software multisim 10.1 sebagai simulator rangkaian, dan software eagle 5.1.1 untuk mendesain jalur rangkaian pada papan PCB. Saat cahaya LASER tidak sampai ke LDR karena terhalang oleh sesuatu, maka rangkaian output yang berupa indikator LED dan panggilan dari telepon rumah akan aktif

  10. Alarm management for process control a best-practice guide for design, implementation, and use of industrial alarm systems

    CERN Document Server

    Rothenberg, Douglas H

    2014-01-01

    No modern industrial enterprise, particularly in such areas as chemical processing, can operate without a secure, and reliable, network of automated monitors and controls. And those operations need alarm systems to alert engineers and managers the moment anything goes wrong or needs attention. This book, by one of the world's leading experts on industrial alarm systems, will provide A to Z coverage of designing, implementing, and maintaining an effective alarm network.

  11. Influência do nível de habilidade e posição corporal no alcance de lactentes Influence of skill level and body position on infants’ reaching

    Directory of Open Access Journals (Sweden)

    RP Carvalho

    2008-06-01

    Full Text Available CONTEXTUALIZAÇÃO: Embora o desenvolvimento do alcance em diferentes posições corporais tem sido amplamente estudado, há poucas pesquisas sobre este assunto, considerando o nível de habilidade do lactente. Objetivo: Verificar como as posições corporais afetam os ajustes proximais e distais do alcance manual de lactentes de quatro a seis meses. MATERIAIS E MÉTODOS: Dez lactentes classificados como menos (n=6 e mais habilidosos (n=4 foram avaliados no mês de aquisição do alcance (M1 e após um mês de prática espontânea (M2, nas posições supina (0º, reclinada (45º e sentada (70º. Foram analisados os ajustes proximais (alcances uni ou bimanuais e distais (mão aberta, semi-aberta ou fechada, posição das mãos no início do movimento (perto ou longe do corpo e movimentos de preensão. RESULTADOS: Houve predomínio de alcances unimanuais para os lactentes mais habilidosos. Os lactentes menos habilidosos apresentaram maior freqüência de alcances unimanuais em M1 (sentado e M2 (supino e sentado. Houve predomínio de alcances com as mãos semi-abertas, exceto para lactentes mais habilidosos em M2 na posição reclinada. Lactentes menos habilidosos apresentaram mãos próximas ao corpo em M1 (reclinado e sentado e M2 (reclinado, enquanto os mais habilidosos iniciaram seus alcances com as mãos longe do corpo em M2 (supino. Lactentes mais habilidosos fizeram mais alcances seguidos de preensão em M2 nas posições supina e sentada. CONCLUSÕES: A posição corporal afeta os ajustes proximais e distais do alcance, de acordo com o nível de habilidade do lactente.BACKGROUND: Although the development of reaching under different body position conditions has been widely studied, little research has addressed this issue considering the infant’s skill level. OBJECTIVE: To investigate how different body positions affect proximal and distal reaching adjustments among four to six-month-old infants. Methods: Ten infants classified as less (n

  12. Reaching High Interactive Levels with Situated Agents

    Directory of Open Access Journals (Sweden)

    Guzmán-Obando J.

    2013-01-01

    Full Text Available El documento presenta un proceso coordinado a través del cual los agentes situados son capaces de trabajar en forma conjunta con el conocimiento de sus habilidades para interactuar con otras entidades similares. Una buena coordinación se deriva de que cada agente situado es capaz de representar sus tasas de idoneidad para llevar a cabo cualquier acción de una manera eficaz. Además de utilizar dichas tasas, un agente puede coordinar sus acciones con otros agentes. En este sentido, cada agente situado trata de seleccionar y realizar sólo las acciones con los mayores índices de ejecución. Algunos experimentos en un verdadero fútbol robótico banco de pruebas están diseñadas para poner en práctica el enfoque coordinado propuesto. Observaciones finales concluir la eficacia y las ventajas de la obra presentada con el fin de aumentar el rendimiento de un equipo integrado por agentes situados cuando tienen que resolver tareas complejas en un escenario dinámico, competitivo e impredecible.

  13. Reaching the unreached.

    Science.gov (United States)

    Ariyaratne, A T

    1989-01-01

    Embodied in the child survival revolution are ideological, methodological, and organizational innovations aimed at radical change in the condition of the world's children as rapidly as possible. In countries such as Sri Lanka, child survival and health for all by the year 2000 often seem to be impossible goals, given the tumultuous socioeconomic and political conditions. In Sri Lanka, the quality of life has been eroded, not enhanced, by the importation of Western technology and managerial capitalism and the destruction of indigenous processes. The chaos and violence that have been brought into the country have made it difficult to reach the poor children, women, and refugees in rural areas with primary health care interventions. Sri Lanka's unreachable--the decision making elites--have blocked access to the unreached--the urban and rural poor. If governments are to reach the unreached, they must remove the obstacles to a people-centered, community development process. It is the people themselves, and the institutions of their creation, that can reach the children amidst them in greatest need. To achieve this task, local communities must be provided with basic human rights, the power to make decisions that affect their lives, necessary resources, and appropriate technologies. Nongovernmental organizations can play a crucial role as bridges between the unreached and the unreachable by promoting community empowerment, aiding in the formation of networks of community organizations, and establishing linkages with government programs. If the ruling elites in developing countries can be persuaded to accommodate the needs and aspirations of those who, to date, have been excluded from the development process, the child survival revolution can be a nonviolent one.

  14. Solar Hydrogen Reaching Maturity

    Directory of Open Access Journals (Sweden)

    Rongé Jan

    2015-09-01

    Full Text Available Increasingly vast research efforts are devoted to the development of materials and processes for solar hydrogen production by light-driven dissociation of water into oxygen and hydrogen. Storage of solar energy in chemical bonds resolves the issues associated with the intermittent nature of sunlight, by decoupling energy generation and consumption. This paper investigates recent advances and prospects in solar hydrogen processes that are reaching market readiness. Future energy scenarios involving solar hydrogen are proposed and a case is made for systems producing hydrogen from water vapor present in air, supported by advanced modeling.

  15. Assessment method to predict the rate of unresolved false alarms

    International Nuclear Information System (INIS)

    Reardon, P.T.; Eggers, R.F.; Heaberlin, S.W.

    1982-06-01

    A method has been developed to predict the rate of unresolved false alarms of material loss in a nuclear facility. The computer program DETRES-1 was developed. The program first assigns the true values of control unit components receipts, shipments, beginning and ending inventories. A normal random number generator is used to generate measured values of each component. A loss estimator is calculated from the control unit's measured values. If the loss estimator triggers a detection alarm, a response is simulated. The response simulation is divided into two phases. The first phase is to simulate remeasurement of the components of the detection loss estimator using the same or better measurement methods or inferences from surrounding control units. If this phase of response continues to indicate a material loss, phase of response simulating a production shutdown and comprehensive cleanout is initiated. A new loss estimator is found, and tested against the alarm thresholds. If the estimator value is below the threshold, the original detection alarm is considered resolved; if above the threshold, an unresolved alarm has occurred. A tally is kept of valid alarms, unresolved false alarms, and failure to alarm upon a true loss

  16. Los Alamos Scientific Laboratory long-range alarm system

    International Nuclear Information System (INIS)

    DesJardin, R.; Machanik, J.

    1980-01-01

    The Los Alamos Scientific Laboratory (LASL) Long-Range Alarm System is described. The last few years have brought significant changes in the Department of Energy regulations for protection of classified documents and special nuclear material. These changes in regulations have forced a complete redesign of the LASL security alarm system. LASL covers many square miles of varying terrain and consists of separate technical areas connected by public roads and communications. A design study over a period of 2 years produced functional specifications for a distributed intelligence, expandable alarm system that will handle 30,000 alarm points from hundreds of data concentrators spread over a 250-km 2 area. Emphasis in the design was on nonstop operation, data security, data communication, and upward expandability to incorporate fire alarms and the computer-aided dispatching of security and fire vehicles. All aspects of the alarm system were to be fault tolerant from the central computer system down to but not including the individual data concentrators. Redundant communications lines travel over public domain from the alarmed area to the central alarm station

  17. Alarm systems a guide to design, management and procurement

    CERN Document Server

    Engineering Equipment and Materials Users' Association. London

    2013-01-01

    Alarm systems form an essential part of the operator interfaces to large modern industrial facilities. They provide vital support to the operators by warning them of situations that need their attention and have an important role in preventing, controlling and mitigating the effects of abnormal situations. Since it was first published in 1999, EEMUA 191 has become the globally accepted and leading guide to good practice for all aspects of alarm systems. The guide, developed by users of alarm systems with input from the GB Health and Safety Executive, gives comprehensive guidance on designing, managing and procuring an effective alarm system. The new Third Edition has been comprehensively updated and includes guidance on implementing the alarm management philosophy in practice; applications in geographically distributed processes; and performance metrics and KPIs.

  18. Development of methods and means to improve a performance of microprocessor shock sensors for car alarms

    Directory of Open Access Journals (Sweden)

    S. A. Vasyukov

    2014-01-01

    -line adjusting characteristic.The developed shock sensor is implemented as stock-produced items included in the car alarm systems of the trademark “Excellent” (made by JSC Magic Ring Ltd.. Operating experience (about 40 thousand pieces realized now allows a proper adjustment of sensors to be possible on cars with any body type. A sufficient level of the sensor sensitivity proves it. The reliability of sensors is confirmed by failures statistics in use (0,08%. Failures are, usually, caused by the SE coil wire break in a soldering point of the coil leads to the printed circuit. However a number of shortcomings are also revealed.Experimental data processing of a test batch of sensors has shown that existing SE of electromagnetic type have the frequency of free vibrations f 805Hz and a decay constant of transient under shock  0,2s. At these parameters a SE pass-band makes only f 1,5Hz, and the sensor responses only to the shock front, rather than to the secondary vibrations of a body. It is an essential drawback of the narrow-band sensors.The experiments conducted on test batch of sensors (10 samples have shown essentially different sensitivities of sensors under shock in the direction of the sensitivity axis and perpendicular to it. This difference reaches 4-6 gradations of sensitivity. It means that when mounting a sensor on the particular car the orientation of sensitivity axis of the sensor will be of crucial importance that is also a shortcoming.It is shown that overcoming the specified drawbacks is hardly because of the improvement of sensitive elements suspension of electromagnetic sensors. Three-axial accelerometers implemented using MEMS technologies seem to be the most perspective application here.

  19. A dynamical management for the handling of alarms at CERN; Une supervision dynamique pour la gestion des alarmes au CERN

    Energy Technology Data Exchange (ETDEWEB)

    Parisot, F

    2007-06-15

    The LHC (large hadron collider) is a huge 27 km long particle accelerator that is being installed 100 meter deep underground near Geneva. This machine requires a sophisticated and complex safety system in charge of the security of staff, infrastructure and equipment. The main features of this system is to be completely redundant by itself and be flexible enough to incorporate new parameters from modification of the accelerating system or from new equipment. The redundancy has been thoroughly studied, for instance a cable can be made of copper threads while the other is an optic fiber which add more safety by being able to cope better with the diversity of accidental situations. The LHC site has been divided into 33 security areas and 4 levels of emergency have been set: level 0 is the conveyance of pure information, level 1 is the conveyance of more important information that requires the intervention of an operator within the day, level 2 requires the intervention of an operator immediately, and level 3 is the emergency level for the intervention of the fire brigade. This safety system is in charge of 300000 parameters from a large variety of sensors and manages 17000 alarms. (A.C.)

  20. BROOKHAVEN: Proton goal reached

    International Nuclear Information System (INIS)

    Anon.

    1995-01-01

    On March 30 the 35-year old Alternating Gradient Synchrotron (AGS) exceeded its updated design goal of 6 x 10 13 protons per pulse (ppp), by accelerating 6.3 x 10 13 ppp, a world record intensity. This goal was set 11 years ago and achieving it called for the construction of a new booster and the reconstruction of much of the AGS. The booster was completed in 1991, and reached its design intensity of 1.5 x 10 13 ppp in 1993. The AGS reconstruction was finished in 1994, and by July of that year the AGS claimed a new US record intensity for a proton synchrotron of 4 x 10 13 ppp, using four booster pulses. Reaching the design intensity was scheduled for 1995. In 1994, the AGS had seemed to be solidly limited to 4 x 10 13 ppp, but in 1995 the operations crew, working on their own in the quiet of the owl shift, steadily improved the intensity, regularly setting new records, much to the bemusement of the machine physicists. The physicists, however, did contribute. A second harmonic radiofrequency cavity in the booster increased the radiofrequency bucket area for capture, raising the booster intensity from 1.7 to 2.1 x 10 13 ppp. In the AGS, new radiofrequency power supplies raised the available voltage from 8 to 13 kV, greatly enhancing the beam loading capabilities of the system. A powerful new transverse damping system successfully controlled instabilities that otherwise would have destroyed the beam in less than a millisecond. Also in the AGS, 35th harmonic octupole resonances were found

  1. BROOKHAVEN: Proton goal reached

    Energy Technology Data Exchange (ETDEWEB)

    Anon.

    1995-09-15

    On March 30 the 35-year old Alternating Gradient Synchrotron (AGS) exceeded its updated design goal of 6 x 10{sup 13} protons per pulse (ppp), by accelerating 6.3 x 10{sup 13} ppp, a world record intensity. This goal was set 11 years ago and achieving it called for the construction of a new booster and the reconstruction of much of the AGS. The booster was completed in 1991, and reached its design intensity of 1.5 x 10{sup 13} ppp in 1993. The AGS reconstruction was finished in 1994, and by July of that year the AGS claimed a new US record intensity for a proton synchrotron of 4 x 10{sup 13} ppp, using four booster pulses. Reaching the design intensity was scheduled for 1995. In 1994, the AGS had seemed to be solidly limited to 4 x 10{sup 13} ppp, but in 1995 the operations crew, working on their own in the quiet of the owl shift, steadily improved the intensity, regularly setting new records, much to the bemusement of the machine physicists. The physicists, however, did contribute. A second harmonic radiofrequency cavity in the booster increased the radiofrequency bucket area for capture, raising the booster intensity from 1.7 to 2.1 x 10{sup 13} ppp. In the AGS, new radiofrequency power supplies raised the available voltage from 8 to 13 kV, greatly enhancing the beam loading capabilities of the system. A powerful new transverse damping system successfully controlled instabilities that otherwise would have destroyed the beam in less than a millisecond. Also in the AGS, 35th harmonic octupole resonances were found.

  2. Changes in Default Alarm Settings and Standard In-Service are Insufficient to Improve Alarm Fatigue in an Intensive Care Unit: A Pilot Project.

    Science.gov (United States)

    Sowan, Azizeh Khaled; Gomez, Tiffany Michelle; Tarriela, Albert Fajardo; Reed, Charles Calhoun; Paper, Bruce Michael

    2016-01-11

    Clinical alarm systems safety is a national concern, specifically in intensive care units (ICUs) where alarm rates are known to be the highest. Interventional projects that examined the effect of changing default alarm settings on overall alarm rate and on clinicians' attitudes and practices toward clinical alarms and alarm fatigue are scarce. To examine if (1) a change in default alarm settings of the cardiac monitors and (2) in-service nursing education on cardiac monitor use in an ICU would result in reducing alarm rate and in improving nurses' attitudes and practices toward clinical alarms. This quality improvement project took place in a 20-bed transplant/cardiac ICU with a total of 39 nurses. We implemented a unit-wide change of default alarm settings involving 17 parameters of the cardiac monitors. All nurses received an in-service education on monitor use. Alarm data were collected from the audit log of the cardiac monitors 10 weeks before and 10 weeks after the change in monitors' parameters. Nurses' attitudes and practices toward clinical alarms were measured using the Healthcare Technology Foundation National Clinical Alarms Survey, pre- and postintervention. Alarm rate was 87.86 alarms/patient day (a total of 64,500 alarms) at the preintervention period compared to 59.18 alarms/patient day (49,319 alarms) postintervention (P=.01). At baseline, Arterial Blood Pressure (ABP), Pair Premature Ventricular Contractions (PVCs), and Peripheral Capillary Oxygen Saturation (SpO2) alarms were the highest. ABP and SpO2 alarms remained among the top three at the postproject period. Out of the 39 ICU nurses, 24 (62%) provided complete pre- and postproject survey questionnaires. Compared to the preintervention survey, no remarkable changes in the postproject period were reported in nurses' attitudes. Themes in the narrative data were related to poor usability of cardiac monitors and the frequent alarms. The data showed great variation among nurses in terms of changing

  3. DTR, Taut Wire System: An alarm barrier with experience

    International Nuclear Information System (INIS)

    Kraft, A.J.

    1986-01-01

    The Taut Wire Fence Alarm System concept was developed and introduced more that fifteen years ago in Israel. A sudden expansion of the nations's border lines, the difficulty to monitor intrusions along those elongated lines and the need for timely as well as accurate armed response to an intrusion attempt dictated the need for an alarming barrier. Traditionally, protection of perimeters was accomplished by the installation of a fence or other type obstacles (man made or natural) and surveillance by manned patrols, fixed observation posts, and/or electronic devices. Defense planners recognized therefore the need for an alarming barrier. A concentrated effort by scientists solved the problem by developing the first Taut Wire Fence Alarm System in a configuration of an alarm barrier. The system was specified to have an extremely low false alarm rate (FAR/NAR), high probability of detection, the capability to follow various terrains, operability in a wide range of environmental conditions, a capability to delay an intruder, ease of installation by unskilled labor, and low maintenance requirements. The authors try here to explain the various constraints and considerations given during the design stages of the Taut Wire Alarm System so as to bring the present magnitude of users to a better understanding of the system's operation

  4. A Nuisance Alarm Data System for evaluation of intrusion detectors

    International Nuclear Information System (INIS)

    Ream, W.K.

    1990-01-01

    A Nuisance Alarm Data System (NADS) was developed to gather long-term background alarm data on exterior intrusion detectors as part of their evaluation. Since nuisance alarms play an important part in the selection of intrusion detectors for use at Department of Energy (DOE) facilities, an economical and reliable way to monitor and record these alarms was needed. NADS consists of an IBM personal computer and printer along with other commercial units to communicate with the detectors, to gather weather data, and to record video for assessment. Each alarm, its assessment, and the weather conditions occurring at alarm time are placed into a data base that is used in the evaluation of the detector. The operating software is written in Turbo Pascal for easy maintenance and modification. A portable system, based on the NADS design, has been built and shipped to other DOE locations to do on-site alarm monitoring. This has been valuable for the comparison of different detectors in the on-site environment and for testing new detectors when the appropriate conditions do not exist or cannot be simulated at the Exterior Intrusion Detection Testbed

  5. Computational Human Performance Modeling For Alarm System Design

    Energy Technology Data Exchange (ETDEWEB)

    Jacques Hugo

    2012-07-01

    The introduction of new technologies like adaptive automation systems and advanced alarms processing and presentation techniques in nuclear power plants is already having an impact on the safety and effectiveness of plant operations and also the role of the control room operator. This impact is expected to escalate dramatically as more and more nuclear power utilities embark on upgrade projects in order to extend the lifetime of their plants. One of the most visible impacts in control rooms will be the need to replace aging alarm systems. Because most of these alarm systems use obsolete technologies, the methods, techniques and tools that were used to design the previous generation of alarm system designs are no longer effective and need to be updated. The same applies to the need to analyze and redefine operators’ alarm handling tasks. In the past, methods for analyzing human tasks and workload have relied on crude, paper-based methods that often lacked traceability. New approaches are needed to allow analysts to model and represent the new concepts of alarm operation and human-system interaction. State-of-the-art task simulation tools are now available that offer a cost-effective and efficient method for examining the effect of operator performance in different conditions and operational scenarios. A discrete event simulation system was used by human factors researchers at the Idaho National Laboratory to develop a generic alarm handling model to examine the effect of operator performance with simulated modern alarm system. It allowed analysts to evaluate alarm generation patterns as well as critical task times and human workload predicted by the system.

  6. Reaching Beyond The Stars

    Science.gov (United States)

    Baker, Mariah; Rosenthal, L.; Gaughan, A.; Hopkins, E.

    2014-01-01

    Strawbridge Observatory at Haverford College is home to a undergraduate-led public observing program. Our program holds ~once monthly public events throughout the academic year that take advantage of eyepiece observing on our 16-inch and 12-inch telescopes as well as of the classroom, library, and projection system. These resources allow us to organize a variety of astronomy related activities that are engaging for individuals of all ages: accessible student talks, current film screenings and even arts and crafts for the families who attend with young children. These events aim to spark curiosity in others about scientific discovery and about the remarkable nature of the world in which we live. In addition to exciting local families about astronomy, this program has excited Haverford students from a range of disciplines about both science and education. Being entirely student led means that we are able to take the initiative in planning, coordinating and running all events, fostering an atmosphere of collaboration, experimentation and commitment amongst our volunteers. Additionally, this program is one of the few at Haverford that regularly reaches beyond the campus walls to promote and build relationships with the outside community. In light of this, our program presents a distinctive and enlightening opportunity for student volunteers: we get to use our scientific backgrounds to educate a general audience, while also learning from them about how to communicate and inspire in others the excitement we feel about the subject of astronomy. The work on this project has been supported by NSF AST-1151462.

  7. GAP-REACH

    Science.gov (United States)

    Lewis-Fernández, Roberto; Raggio, Greer A.; Gorritz, Magdaliz; Duan, Naihua; Marcus, Sue; Cabassa, Leopoldo J.; Humensky, Jennifer; Becker, Anne E.; Alarcón, Renato D.; Oquendo, María A.; Hansen, Helena; Like, Robert C.; Weiss, Mitchell; Desai, Prakash N.; Jacobsen, Frederick M.; Foulks, Edward F.; Primm, Annelle; Lu, Francis; Kopelowicz, Alex; Hinton, Ladson; Hinton, Devon E.

    2015-01-01

    Growing awareness of health and health care disparities highlights the importance of including information about race, ethnicity, and culture (REC) in health research. Reporting of REC factors in research publications, however, is notoriously imprecise and unsystematic. This article describes the development of a checklist to assess the comprehensiveness and the applicability of REC factor reporting in psychiatric research publications. The 16-itemGAP-REACH© checklist was developed through a rigorous process of expert consensus, empirical content analysis in a sample of publications (N = 1205), and interrater reliability (IRR) assessment (N = 30). The items assess each section in the conventional structure of a health research article. Data from the assessment may be considered on an item-by-item basis or as a total score ranging from 0% to 100%. The final checklist has excellent IRR (κ = 0.91). The GAP-REACH may be used by multiple research stakeholders to assess the scope of REC reporting in a research article. PMID:24080673

  8. UX-15 Reaches LEP

    CERN Multimedia

    2001-01-01

    The creation of the world's largest sandstone cavern, not a small feat! At the bottom, cave-in preventing steel mesh can be seen clinging to the top of the tunnel. The digging of UX-15, the cavern that will house ATLAS, reached the upper ceiling of LEP on October 10th. The breakthrough which took place nearly 100 metres underground occurred precisely on schedule and exactly as planned. But much caution was taken beforehand to make the LEP breakthrough clean and safe. To prevent the possibility of cave-ins in the side tunnels that will eventually be attached to the completed UX-15 cavern, reinforcing steel mesh was fixed into the walls with bolts. Obviously no people were allowed in the LEP tunnels below UX-15 as the breakthrough occurred. The area was completely evacuated and fences were put into place to keep all personnel out. However, while personnel were being kept out of the tunnels below, this has been anything but the case for the work taking place up above. With the creation of the world's largest...

  9. Development and Characteristics of an ANSI/ANS-8-3-1997-Compliant Criticality Alarm

    International Nuclear Information System (INIS)

    McMahan, J.W.

    2002-01-01

    A Criticality Alarm System (CAS) was developed at the Savannah River Site (SRS) beginning in 1964, which used an ionization chamber and battery - powered tube amplifier. Three generations of improvements are described here, to improve reliability, reduce maintenance cost, and eliminate false alarms. The original configuration for the CAS instrument (called Nuclear Incident Monitor-NIM model I) was a single chassis containing all components including the batteries and the detector, and a warning bell and light. To decrease the contamination potential of NIMs, the detector can now be removed from the chassis and remotely located for some installations. In the presently used second generation model II, each unit consists of an ion chamber with electrometer amplifier, voltage comparator circuits, battery and power supplies, monitoring circuits, and relay alarm circuits. The electrometer circuit is housed with the ion chamber in an aluminum enclosure. The remaining components are located in the main chassis. An alarm point of one R per Hour is adjustable over a narrow range for calibration purposes. Evacuation zones were determined on a 25 rad dose basis, and then modified to the 12 rad level

  10. New Concept For Alarm Structure And Management In Dcs Systems

    Directory of Open Access Journals (Sweden)

    Mohammed Hegazy

    2015-08-01

    Full Text Available The objective of this paper is to set new standard for good design and best practice to applied when any DCS ManufacturesSuppliers configure process alarm system in any oil refining oil and gas production gas-handling facilities gasification plant or any chemical processing plant and thereby to optimizeminimize unnecessary alarms from reporting to operator workstations CAD Control Alarm Display. These views based on the experience acquired and implemented during involvement with the commissioning and startup of two DCS projects in Mina Al-Ahmadi Refinery Kuwait.

  11. Design of SMART alarm system using main memory database

    International Nuclear Information System (INIS)

    Jang, Kue Sook; Seo, Yong Seok; Park, Keun Oak; Lee, Jong Bok; Kim, Dong Hoon

    2001-01-01

    To achieve design goal of SMART alarm system, first of all we have to decide on how to handle and manage alarm information and how to use database. So this paper analyses concepts and deficiencies of main memory database applied in real time system. And this paper sets up structure and processing principles of main memory database using nonvolatile memory such as flash memory and develops recovery strategy and process board structures using these. Therefore this paper shows design of SMART alarm system is suited functions and requirements

  12. Advanced pulse oximeter signal processing technology compared to simple averaging. II. Effect on frequency of alarms in the postanesthesia care unit.

    Science.gov (United States)

    Rheineck-Leyssius, A T; Kalkman, C J

    1999-05-01

    To determine the effect of a new pulse oximeter (Nellcor Symphony N-3000, Pleasanton, CA) with signal processing technique (Oxismart) on the incidence of false alarms in the postanesthesia care unit (PACU). Prospective study. Nonuniversity hospital. 603 consecutive ASA physical status I, II, and III patients recovering from general or regional anesthesia in the PACU. We compared the number of alarms produced by a recently developed "third"-generation pulse oximeter (Nellcor Symphony N-3000) with Oxismart signal processing technique and a conventional pulse oximeter (Criticare 504, Waukesha, WI). Patients were randomly assigned to either a Nellcor pulse oximeter or a Criticare with the signal averaging time set at either 12 or 21 seconds. For each patient the number of false (artifact) alarms was counted. The Nellcor generated one false alarm in 199 patients and 36 (in 31 patients) "loss of pulse" alarms. The conventional pulse oximeter with the averaging time set at 12 seconds generated a total of 32 false alarms in 17 of 197 patients [compared with the Nellcor, relative risk (RR) 0.06, confidence interval (CI) 0.01 to 0.25] and a total of 172 "loss of pulse" alarms in 79 patients (RR 0.39, CI 0.28 to 0.55). The conventional pulse oximeter with the averaging time set at 21 seconds generated 12 false alarms in 11 of 207 patients (compared with the Nellcor, RR 0.09, CI 0.02 to 0.48) and a total of 204 "loss of pulse" alarms in 81 patients (RR 0.40, CI 0.28 to 0.56). The lower incidence of false alarms of the conventional pulse oximeter with the longest averaging time compared with the shorter averaging time did not reach statistical significance (false alarms RR 0.62, CI 0.3 to 1.27; "loss of pulse" alarms RR 0.98, CI 0.77 to 1.3). To date, this is the first report of a pulse oximeter that produced almost no false alarms in the PACU.

  13. A new method for defining and managing process alarms and for correcting process operation when an alarm occurs

    International Nuclear Information System (INIS)

    Brooks, Robin; Thorpe, Richard; Wilson, John

    2004-01-01

    A new mathematical treatment of alarms that considers them as multi-variable interactions between process variables has provided the first-ever method to calculate values for alarm limits. This has resulted in substantial reductions in false alarms and hence in alarm annunciation rates in field trials. It has also unified alarm management, process control and product quality control into a single mathematical framework so that operations improvement and hence economic benefits are obtained at the same time as increased process safety. Additionally, an algorithm has been developed that advises what changes should be made to Manipulable process variables to clear an alarm. The multi-variable Best Operating Zone at the heart of the method is derived from existing historical data using equation-free methods. It does not require a first-principles process model or an expensive series of process identification experiments. Integral with the method is a new format Process Operator Display that uses only existing variables to fully describe the multi-variable operating space. This combination of features makes it an affordable and maintainable solution for small plants and single items of equipment as well as for the largest plants. In many cases, it also provides the justification for the investments about to be made or already made in process historian systems. Field Trials have been and are being conducted at IneosChlor and Mallinckrodt Chemicals, both in the UK, of the new geometric process control (GPC) method for improving the quality of both process operations and product by providing Process Alarms and Alerts of much high quality than ever before. The paper describes the methods used, including a simple visual method for Alarm Rationalisation that quickly delivers large sets of Consistent Alarm Limits, and the extension to full Alert Management with highlights from the Field Trials to indicate the overall effectiveness of the method in practice

  14. A new method for defining and managing process alarms and for correcting process operation when an alarm occurs

    Energy Technology Data Exchange (ETDEWEB)

    Brooks, Robin [Curvaceous Software Limited, P.O. Box 43, Gerrards Cross, Bucks SL98UX (United Kingdom)]. E-mail: enquiries@curvaceous.com; Thorpe, Richard [Curvaceous Software Limited, P.O. Box 43, Gerrards Cross, Bucks SL98UX (United Kingdom); Wilson, John [Curvaceous Software Limited, P.O. Box 43, Gerrards Cross, Bucks SL98UX (United Kingdom)

    2004-11-11

    A new mathematical treatment of alarms that considers them as multi-variable interactions between process variables has provided the first-ever method to calculate values for alarm limits. This has resulted in substantial reductions in false alarms and hence in alarm annunciation rates in field trials. It has also unified alarm management, process control and product quality control into a single mathematical framework so that operations improvement and hence economic benefits are obtained at the same time as increased process safety. Additionally, an algorithm has been developed that advises what changes should be made to Manipulable process variables to clear an alarm. The multi-variable Best Operating Zone at the heart of the method is derived from existing historical data using equation-free methods. It does not require a first-principles process model or an expensive series of process identification experiments. Integral with the method is a new format Process Operator Display that uses only existing variables to fully describe the multi-variable operating space. This combination of features makes it an affordable and maintainable solution for small plants and single items of equipment as well as for the largest plants. In many cases, it also provides the justification for the investments about to be made or already made in process historian systems. Field Trials have been and are being conducted at IneosChlor and Mallinckrodt Chemicals, both in the UK, of the new geometric process control (GPC) method for improving the quality of both process operations and product by providing Process Alarms and Alerts of much high quality than ever before. The paper describes the methods used, including a simple visual method for Alarm Rationalisation that quickly delivers large sets of Consistent Alarm Limits, and the extension to full Alert Management with highlights from the Field Trials to indicate the overall effectiveness of the method in practice.

  15. A new method for defining and managing process alarms and for correcting process operation when an alarm occurs.

    Science.gov (United States)

    Brooks, Robin; Thorpe, Richard; Wilson, John

    2004-11-11

    A new mathematical treatment of alarms that considers them as multi-variable interactions between process variables has provided the first-ever method to calculate values for alarm limits. This has resulted in substantial reductions in false alarms and hence in alarm annunciation rates in field trials. It has also unified alarm management, process control and product quality control into a single mathematical framework so that operations improvement and hence economic benefits are obtained at the same time as increased process safety. Additionally, an algorithm has been developed that advises what changes should be made to Manipulable process variables to clear an alarm. The multi-variable Best Operating Zone at the heart of the method is derived from existing historical data using equation-free methods. It does not require a first-principles process model or an expensive series of process identification experiments. Integral with the method is a new format Process Operator Display that uses only existing variables to fully describe the multi-variable operating space. This combination of features makes it an affordable and maintainable solution for small plants and single items of equipment as well as for the largest plants. In many cases, it also provides the justification for the investments about to be made or already made in process historian systems. Field Trials have been and are being conducted at IneosChlor and Mallinckrodt Chemicals, both in the UK, of the new geometric process control (GPC) method for improving the quality of both process operations and product by providing Process Alarms and Alerts of much high quality than ever before. The paper describes the methods used, including a simple visual method for Alarm Rationalisation that quickly delivers large sets of Consistent Alarm Limits, and the extension to full Alert Management with highlights from the Field Trials to indicate the overall effectiveness of the method in practice.

  16. Analysis of Statistical Distributions Used for Modeling Reliability and Failure Rate of Temperature Alarm Circuit

    International Nuclear Information System (INIS)

    EI-Shanshoury, G.I.

    2011-01-01

    Several statistical distributions are used to model various reliability and maintainability parameters. The applied distribution depends on the' nature of the data being analyzed. The presented paper deals with analysis of some statistical distributions used in reliability to reach the best fit of distribution analysis. The calculations rely on circuit quantity parameters obtained by using Relex 2009 computer program. The statistical analysis of ten different distributions indicated that Weibull distribution gives the best fit distribution for modeling the reliability of the data set of Temperature Alarm Circuit (TAC). However, the Exponential distribution is found to be the best fit distribution for modeling the failure rate

  17. Changes in 900 MW PWR alarm processing policy

    Energy Technology Data Exchange (ETDEWEB)

    Pont, M [Electricite de France, Generation and Transmission, Nuclear Power Plant Operations, Paris (France)

    1997-09-01

    Following a brief description of the current 900 MW PWR alarm processing system, this document presents the feasibility study carried out within the scope of the Instrumentation and Control Refurbishment project (R2C). (author). 4 figs, tabs.

  18. Changes in 900 MW PWR alarm processing policy

    International Nuclear Information System (INIS)

    Pont, M.

    1997-01-01

    Following a brief description of the current 900 MW PWR alarm processing system, this document presents the feasibility study carried out within the scope of the Instrumentation and Control Refurbishment project (R2C). (author). 4 figs, tabs

  19. rf duress alarms: market survey and preliminary characterization

    International Nuclear Information System (INIS)

    Draper, B.L.

    1979-05-01

    This report represents the first phase of the duress alarm studies. Presented here are the results of an extensive market survey and some preliminary observations on the effectiveness of many system components

  20. Ship cabin leakage alarm based on ARM SCM

    Science.gov (United States)

    Qu, Liyan

    2018-03-01

    If there is a leakage in the cabin of a sailing ship, it is a major accident that threatens the personnel and property of the ship. If we can’t take timely measures, there will be a devastating disaster. In order to judge the leakage of the cabin, it is necessary to set up a leakage alarm system, so as to achieve the purpose of detecting and alarming the leakage of the cabin, and avoid the occurrence of accidents. This paper discusses the design of ship cabin leakage alarm system based on ARM SCM. In order to ensure the stability and precision of the product, the hardware design of the alarm system is carried out, such as circuit design, software design, the programming of SCM, the software programming of upper computer, etc. It is hoped that it can be of reference value to interested readers.

  1. A weighted dissimilarity index to isolate faults during alarm floods

    CERN Document Server

    Charbonnier, S; Gayet, P

    2015-01-01

    A fault-isolation method based on pattern matching using the alarm lists raised by the SCADA system during an alarm flood is proposed. A training set composed of faults is used to create fault templates. Alarm vectors generated by unknown faults are classified by comparing them with the fault templates using an original weighted dissimilarity index that increases the influence of the few alarms relevant to diagnose the fault. Different decision strategies are proposed to support the operator in his decision making. The performances are evaluated on two sets of data: an artificial set and a set obtained from a highly realistic simulator of the CERN Large Hadron Collider process connected to the real CERN SCADA system.

  2. 46 CFR 97.37-50 - Ventilation alarm failure.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 4 2010-10-01 2010-10-01 false Ventilation alarm failure. 97.37-50 Section 97.37-50... OPERATIONS Markings for Fire and Emergency Equipment, Etc. § 97.37-50 Ventilation alarm failure. (a) The...-inch letters “VENTILATION FAILURE IN VEHICULAR SPACE.” (b) [Reserved] [CGFR 66-33, 31 FR 15286, Dec. 6...

  3. [Citalopram, escitalopram and prolonged QT: warning or alarm?].

    Science.gov (United States)

    Alvarez, Enric; Vieira, Sara; Garcia-Moll, Xavier

    2014-01-01

    The alerts issued by regulatory agencies on the potential cardiac toxicity of citalopram and escitalopram have caused alarm among clinicians. A review of the data concerning this topic shows that the alarm should be limited to patients with a history of syncope or poisoning. As a precautionary measure, an electrocardiogram should be performed on elderly patients. Copyright © 2013 SEP y SEPB. Published by Elsevier España. All rights reserved.

  4. A study of reset mode in advanced alarm system simulator

    International Nuclear Information System (INIS)

    Yenn, T. C.; Hwang, S. L.; Huang, F. H.; Yu, A. C.; Hsu, C. C.; Huang, H. W.

    2006-01-01

    An automation function has been widely applied in main control room of nuclear power plants. That leads to a new issue of human-automation interaction, which considers human operational performance in automated systems. In this research is the automation alarm reset in the advanced alarm system (AAS) of Advanced Nuclear Power Plant in Taiwan. Since alarms are very crucial for the understanding of the status of the plant as well as the reset function of alarm system will be changed from fully manual to fully automatic, it is very important to test and evaluate the performance and the effect of reset modes in AAS. The purpose of this paper is to evaluate the impact of the auto-reset alarm system on the plant performance and on operators' preference and task load. To develop a dynamic simulator as an AAS was conducted to compare manual and automatic reset function of alarm system on task performance and subjective ratings of task workload, comprehension, and preference. The simulation includes PCTRAN model and alarm software processing. The final results revealed that, using the auto-reset mode, participants had lower task load index (TLX) on effort in the first test trial and was more satisfied in multiple tasks condition. In contrast, using manual reset mode, participants were more satisfied on alarm handling, monitoring, and decision making. In other words, either reset mode in the study has unique features to assist operator, but is insufficient. The reset function in AAS therefore should be very flexible. Additionally, the experimental results also pointed out that the user interfaces need to be improved. Those experiences will be helpful for human factors verification and validation in the near future. (authors)

  5. Development of alarm handling methods for boiling water reactors

    Energy Technology Data Exchange (ETDEWEB)

    Yukiharu, Ohga; Hiroshi, Seki; Setsuo, Arita [Power and Industrial Systems R and D Div., Hitachi, Ltd., Hitachi, Ibaraki (Japan)

    1997-09-01

    A method was developed to select important alarms in two steps: first, selection is based on the physical relationship between the alarms, and second, selection is according to the initial event. An approach combining a neural network and knowledge processing was proposed to identify the event rapidly. A prototype system was evaluated in the Kashiwazaki/Kariwa-4 Nuclear Power Plant during the startup test. The evaluation test confirmed that about 30% of the alarms are selected from among the many activated alarms. The second method, dealing with presentation, supports operators in their selection and confirmation of the required information for plant operation. The method selects and offers plant information in response to plant status changes and operators` demands. The selection procedure is based on the knowledge and data as structured by the plant functional structure; i.e. a means-ends abstraction hierarchy model. A prototype system was evaluated using a BWR simulator. The results showed that appropriate information items are automatically selected according to plant status changes and information on generated alarms is presented to operators together with the related trend graph and system diagram. Answers are generated in reply to the operators` demands and operators can confirm the generated alarms on each plant function, such as systems and components. 8 refs, 10 figs, 2 tabs.

  6. Measurement of the Portsmouth Gaseous Diffusion Plant criticality accident alarm

    International Nuclear Information System (INIS)

    Tayloe, R.W. Jr.; D'Aquila, D.M.; McGinnis, R.B.

    1991-01-01

    The nuclear criticality accident radiation alarm system installed at the Portsmouth Gaseous Diffusion Plant was tested extensively at critical facilities located at the Los Alamos National Laboratory. The ability of the neutron scintillator radiation detection units to respond to a minimum accident of concern as defined in Standard ANSI/ANS-83.-1986 was demonstrated. Detector placement and the established trip point are based on shielding calculations performed by the Oak Ridge National Laboratory and criticality specialists at the Portsmouth plant. Based on these experiments and calculations, a detector trip point of 5 mrad/h in air is used. Any credible criticality accident is expected to produce neutron radiation fields >5 mrad/h in air at one or more radiation alarm locations. Each radiation alarm location has a cluster of three detectors that employs a two-out-of-three alarm logic. Earlier work focused on testing the alarm logic latching circuitry. This work was directed toward measurements involving the actual audible alarm signal delivered

  7. Do aphid colonies amplify their emission of alarm pheromone?

    Science.gov (United States)

    Hatano, Eduardo; Kunert, Grit; Bartram, Stefan; Boland, Wilhelm; Gershenzon, Jonathan; Weisser, Wolfgang W

    2008-09-01

    When aphids are attacked by natural enemies, they emit alarm pheromone to alert conspecifics. For most aphids tested, (E)-beta-farnesene (EBF) is the main, or only, constituent of the alarm pheromone. In response to alarm pheromone, alerted aphids drop off the plant, walk away, or attempt to elude predators. However, under natural conditions, EBF concentration might be low due to the low amounts emitted, to rapid air movement, or to oxidative degradation. To ensure that conspecifics are warned, aphids might conceivably amplify the alarm signal by emitting EBF in response to EBF emitted by other aphids. To examine whether such amplification occurs, we synthesized deuterated EBF (DEBF), which allowed us to differentiate between applied and aphid-derived chemical. Colonies of Acyrthosiphon pisum were treated with DEBF, and headspace volatiles were collected and analyzed for evidence of aphid-derived EBF. No aphid-derived EBF was detected, suggesting that amplification of the alarm signal does not occur. We discuss the disadvantages of alarm signal reinforcement.

  8. Development of alarm handling methods for boiling water reactors

    International Nuclear Information System (INIS)

    Ohga Yukiharu; Seki Hiroshi; Arita Setsuo

    1997-01-01

    A method was developed to select important alarms in two steps: first, selection is based on the physical relationship between the alarms, and second, selection is according to the initial event. An approach combining a neural network and knowledge processing was proposed to identify the event rapidly. A prototype system was evaluated in the Kashiwazaki/Kariwa-4 Nuclear Power Plant during the startup test. The evaluation test confirmed that about 30% of the alarms are selected from among the many activated alarms. The second method, dealing with presentation, supports operators in their selection and confirmation of the required information for plant operation. The method selects and offers plant information in response to plant status changes and operators' demands. The selection procedure is based on the knowledge and data as structured by the plant functional structure; i.e. a means-ends abstraction hierarchy model. A prototype system was evaluated using a BWR simulator. The results showed that appropriate information items are automatically selected according to plant status changes and information on generated alarms is presented to operators together with the related trend graph and system diagram. Answers are generated in reply to the operators' demands and operators can confirm the generated alarms on each plant function, such as systems and components. 8 refs, 10 figs, 2 tabs

  9. Alerting the apathetic and reassuring the alarmed: communicating about radon risk in three communities. Final report

    International Nuclear Information System (INIS)

    Chess, C.; Hance, B.J.

    1988-01-01

    Public reaction to the risk from radon varied widely in three communities chosen for qualitative analysis. In Boyertown, PA, some residents were very alarmed, but most were apathetic toward this newly identified environmental risk. In Clinton, NJ, residents were concerned and worked with the mayor and the state to determine whether they were at risk and to disseminate information about mitigation of high indoor radon levels. Residents in Vernon, New Jersey were very alarmed and actively opposed the state's decision to site low-level radium wastes there. The qualitative study examines why reactions differed among the three communities, and extracts lessons for communicating about the risk from radon. These lessons should apply to communicating about other environmental hazards to individuals and communities

  10. Guideline for the licensing of various designs of ionisation smoke alarms

    International Nuclear Information System (INIS)

    1978-01-01

    The purpose of the guideline is to set up criteria for the construction licensing of ionisation smoke alarms and to determine their use. This guarantees that only those ISA are licensed which correspond to the level of science and technology and that the ISA to be licensed fulfills the demands of radiation protection. The recommendations of OECD/NEA are also born in mind. (orig./HP) [de

  11. Simulator testing of the Westinghouse aware alarm management system

    Energy Technology Data Exchange (ETDEWEB)

    Carrera, J P; Easter, J R; Roth, E M [Westinghouse Electric Corp., Pittsburgh, PA (United States)

    1997-09-01

    Over the last year, Westinghouse engineers and operators from the Beznau nuclear power station (KKB), owned by the Nordostschweizerische Krafwerke AG of Baden, Switzerland, have been installing and testing the Westinghouse AWARE Alarm Management System in Beznau/SNUPPS operator training simulator, owned and operated by the Westinghouse Electric Corp., in Waltz Mill, PA, USA. The testing has focused primarily on validating the trigger logic data base and on familiarizing the utility`s training department with the operation of the system in a real-time environment. Some of the tests have included plant process scenarios in which the computerized Emergency Procedures were available and used through the COMPRO (COMputerized PROcedures) System in conjunction with the AWARE System. While the results to date are qualitative from the perspective of system performance and improvement in message presentation, the tests have generally confirmed the expectations of the design. There is a large reduction in the number of messages that the control room staff must deal with during major process abnormalities, yet at times of relative minor disturbances, some additional messages are available which add clarification, e.g., ``Pump Trouble`` messages. The ``flow`` of an abnormality as it progresses from one part of the plant`s processes to another is quite visible. Timing of the messages and the lack of message avalanching is proving to give the operators additional time to respond to messages. Generally, the anxiety level to ``do something`` immediately upon a reactor trip appears to be reduced. (author). 8 refs.

  12. Simulator testing of the Westinghouse aware alarm management system

    International Nuclear Information System (INIS)

    Carrera, J.P.; Easter, J.R.; Roth, E.M.

    1997-01-01

    Over the last year, Westinghouse engineers and operators from the Beznau nuclear power station (KKB), owned by the Nordostschweizerische Krafwerke AG of Baden, Switzerland, have been installing and testing the Westinghouse AWARE Alarm Management System in Beznau/SNUPPS operator training simulator, owned and operated by the Westinghouse Electric Corp., in Waltz Mill, PA, USA. The testing has focused primarily on validating the trigger logic data base and on familiarizing the utility's training department with the operation of the system in a real-time environment. Some of the tests have included plant process scenarios in which the computerized Emergency Procedures were available and used through the COMPRO (COMputerized PROcedures) System in conjunction with the AWARE System. While the results to date are qualitative from the perspective of system performance and improvement in message presentation, the tests have generally confirmed the expectations of the design. There is a large reduction in the number of messages that the control room staff must deal with during major process abnormalities, yet at times of relative minor disturbances, some additional messages are available which add clarification, e.g., ''Pump Trouble'' messages. The ''flow'' of an abnormality as it progresses from one part of the plant's processes to another is quite visible. Timing of the messages and the lack of message avalanching is proving to give the operators additional time to respond to messages. Generally, the anxiety level to ''do something'' immediately upon a reactor trip appears to be reduced. (author). 8 refs

  13. Attributions of cancer 'alarm' symptoms in a community sample.

    Directory of Open Access Journals (Sweden)

    Katriina L Whitaker

    Full Text Available Attribution of early cancer symptoms to a non-serious cause may lead to longer diagnostic intervals. We investigated attributions of potential cancer 'alarm' and non-alarm symptoms experienced in everyday life in a community sample of adults, without mention of a cancer context.A questionnaire was mailed to 4858 adults (≥50 years old, no cancer diagnosis through primary care, asking about symptom experiences in the past 3 months. The word cancer was not mentioned. Target 'alarm' symptoms, publicised by Cancer Research UK, were embedded in a longer symptom list. For each symptom experienced, respondents were asked for their attribution ('what do you think caused it', concern about seriousness ('not at all' to 'extremely', and help-seeking ('did you contact a doctor about it': Yes/No.The response rate was 35% (n = 1724. Over half the respondents (915/1724; 53% had experienced an 'alarm' symptom, and 20 (2% cited cancer as a possible cause. Cancer attributions were highest for 'unexplained lump'; 7% (6/87. Cancer attributions were lowest for 'unexplained weight loss' (0/47. A higher proportion (375/1638; 23% were concerned their symptom might be 'serious', ranging from 12% (13/112 for change in a mole to 41% (100/247 for unexplained pain. Just over half had contacted their doctor about their symptom (59%, although this varied by symptom. Alarm symptoms were appraised as more serious than non-alarm symptoms, and were more likely to trigger help-seeking.Consistent with retrospective reports from cancer patients, 'alarm' symptoms experienced in daily life were rarely attributed to cancer. These results have implications for understanding how people appraise and act on symptoms that could be early warning signs of cancer.

  14. Alarm pheromone processing in the ant brain: an evolutionary perspective

    Directory of Open Access Journals (Sweden)

    Makoto Mizunami

    2010-06-01

    Full Text Available Social insects exhibit sophisticated communication by means of pheromones, one example of which is the use of alarm pheromones to alert nestmates for colony defense. We review recent advances in the understanding of the processing of alarm pheromone information in the ant brain. We found that information about formic acid and n-undecane, alarm pheromone components, is processed in a set of specific glomeruli in the antennal lobe of the ant Camponotus obscuripes. Alarm pheromone information is then transmitted, via projection neurons, to the lateral horn and the calyces of the mushroom body of the protocerebrum. In the lateral horn, we found a specific area where terminal boutons of alarm pheromone-sensitive projection neurons are more densely distributed than in the rest of the lateral horn. Some neurons in the protocerebrum responded specifically to formic acid or n-undecane and they may participate in the control of behavioral responses to each pheromone component. Other neurons, especially those originating from the mushroom body lobe, responded also to non-pheromonal odors and may play roles in integration of pheromonal and non-pheromonal signals. We found that a class of neurons receive inputs in the lateral horn and the mushroom body lobe and terminate in a variety of premotor areas. These neurons may participate in the control of aggressive behavior, which is sensitized by alarm pheromones and is triggered by non-pheromonal sensory stimuli associated with a potential enemy. We propose that the alarm pheromone processing system has evolved by differentiation of a part of general odor processing system.

  15. Reach the sky

    Science.gov (United States)

    Mariana Peicuti, Cristina

    2017-04-01

    I am working as primary teacher at Scoala Gimnaziala Dumbrava,Timis County, Romania & my pupils has 6 to 10 years old. I was&I am a main pillar in my community, always disseminating knowledge and experience to students, other teachers in the school area &Timis County.Astronomy is the must favorite subject of my students from my classes. They are very courious & always come to me with questions about Earth and Sky because Curriculum scientific disciplines provides too little information about Earth and Sky.I need to know more about how to teach space contents into my classes&what competencies can form in elementary school and also to share my experience to the others.As a result of participation at this meeting I want to attract as many students to astronomy,science/STEM disciplines&space technologies, to astronomy topics and exploration of outer space.Schools needs to be prepared for social life needs,new generations needs,on science/space technologies,which are one of the key points for developing the knowledge society.I intend to introduce new scientific activities as part of the existing curriculum.I am passionate about astronomy,I need to know new approaches and new ideas for primary because I think Science is very important in daily life. Here are some developed activities with pupils from K-2 grade levels wich I wish share with colleagues in Viena. Subject: MATHEMATICS. Primary Topic: MEASUREMENT : -+= ☼ Rockets by Size. Students cut out,color and sequence paper rockets/Read the information on the International Space Station and rockets/Gather pictures of different types of rockets/Print/cut out/color&laminate rocket drawings/Find objects in the room to put in order by height. ☼ Oil Spot Photometer - Measure the brightness of the sun using cooking oil and a white card. A smear of oil on a white card becomes a powerful tool for comparing the brightness of two light sources, including the sun. ☼ The Sundial & Making Shadows-device to measure time by the

  16. Alarm reduction with correlation analysis; Larmsanering genom korrelationsanalys

    Energy Technology Data Exchange (ETDEWEB)

    Bergquist, Tord; Ahnlund, Jonas; Johansson, Bjoern; Gaardman, Lennart; Raaberg, Martin [Lund Univ. (Sweden). Dept. of Information Technology

    2004-09-01

    This project's main interest is to improve the overall alarm situation in the control rooms. By doing so, the operators working environment is less overstrained, which simplifies the decision-making. According to a study of the British refinery industry, the operators make wrong decisions in four times out of ten due to badly tuned alarm systems, with heavy expenses as a result. Furthermore, a more efficiently alarm handling is estimated to decrease the production loss with between three and eight percent. This sounds, according to Swedish standards, maybe a bit extreme, but there is no doubt about the benefits of having a well-tuned alarm system. This project can be seen as an extension of 'General Methods for Alarm Reduction' (VARMEFORSK--835), where the process improvements were the result of suggestions tailored for every signal. Here, instead causal dependences in the process are examined. A method for this, specially designed to fit process signals, has been developed. It is called MLPC (Multiple Local Property Correlation) and could be seen as an unprejudiced way of increase the information value in the process. There are a number of ways to make use of the additional process understanding a correlation analysis provides. In the report some are mentioned, foremost aiming to improve the alarm situation for operators. Signals from two heating plants have been analyzed with MLPC. In simulations, with the use of the result from these analyses as a base, a large number of alarms have been successfully suppressed. The results have been studied by personal with process knowledge, and they are very positive to the use of MLPC and they express many benefits by the clarification of process relations. It was established in 'General Methods for Alarm Reduction' that low pass filter are superior to mean value filter and time delay when trying to suppress alarms. As a result, a module for signal processing has been developed. The main purpose is

  17. Nuclear-power-plant perimeter-intrusion alarm systems

    International Nuclear Information System (INIS)

    Halsey, D.J.

    1982-04-01

    Timely intercept of an intruder requires the examination of perimeter barriers and sensors in terms of reliable detection, immediate assessment and prompt response provisions. Perimeter security equipment and operations must at the same time meet the requirements of the Code of Federal Regulations, 10 CFR 73.55 with some attention to the performance and testing figures of Nuclear Regulatory Guide 5.44, Revision 2, May 1980. A baseline system is defined which recommends a general approach to implementing perimeter security elements: barriers, lighting, intrusion detection, alarm assessment. The baseline approach emphasizes cost/effectiveness achieved by detector layering and logic processing of alarm signals to produce reliable alarms and low nuisance alarm rates. A cost benefit of layering along with video assessment is reduction in operating expense. The concept of layering is also shown to minimize testing costs where detectability performance as suggested by Regulatory Guide 5.44 is to be performed. Synthesis of the perimeter intrusion alarm system and limited testing of CCTV and Video Motion Detectors (VMD), were performed at E-Systems, Greenville Division, Greenville, Texas during 1981

  18. Monitoring techniques and alarm procedures for CMS services and sites in WLCG

    CERN Document Server

    Molina-Perez, Jorge Amando

    2012-01-01

    The CMS offline computing system is composed of roughly 80 sites (including most experienced T3s) and a number of central services to distribute, process and analyze data worldwide. A high level of stability and reliability is required from the underlying infrastructure and services, partially covered by local or automated monitoring and alarming systems such as Lemon and SLS; the former collects metrics from sensors installed on computing nodes and triggers alarms when values are out of range, the latter measures the quality of service and warns managers when service is affected. CMS has established computing shift procedures with personnel operating worldwide from remote Computing Centers, under the supervision of the Computing Run Coordinator on duty at CERN. This dedicated 24/7 computing shift personnel is contributing to detect and react timely on any unexpected error and hence ensure that CMS workflows are carried out efficiently and in a sustained manner. Synergy among all the involved actors is explo...

  19. Development of a criticality alarm system neutron detector: Final project report

    International Nuclear Information System (INIS)

    O'Dell, A.A.

    1989-05-01

    The primary objective of this project was to develop a prototype neutron detector for use in criticality alarm systems (CASs) at US Department of Energy (DOE) and contractor facilities wherever significant amounts of fissile material are processed or stored. Constraints placed on the design of the detector were that the overall size of the detector was to be as small as practical, the input voltage requirements were to be no more than 24 V, and that the gamma sensitivity would be as low as possible. Also, the detector should give dosimetric neutron response, and should have sufficient temporal capabilities to measure the entire range from fast (>1 ms) to slow (seconds to minutes) excursions, and sufficient dynamic range to measure from background to over 100 times background levels to insure proper activation of the Immediate Evacuation Alarm (IEA). Finally, the detector should insure rapid (<1 s) activation of the IEA in the event of a criticality excursion. 24 figs., 11 tabs

  20. FAULT DIAGNOSIS WITH MULTI-STATE ALARMS IN A NUCLEAR POWER CONTROL SIMULATOR

    Energy Technology Data Exchange (ETDEWEB)

    Austin Ragsdale; Roger Lew; Brian P. Dyre; Ronald L. Boring

    2012-10-01

    This research addresses how alarm systems can increase operator performance within nuclear power plant operations. The experiment examined the effect of two types of alarm systems (two-state and three-state alarms) on alarm compliance and diagnosis for two types of faults differing in complexity. We hypothesized three-state alarms would improve performance in alarm recognition and fault diagnoses over that of two-state alarms. We used sensitivity and criterion based on Signal Detection Theory to measure performance. We further hypothesized that operator trust would be highest when using three-state alarms. The findings from this research showed participants performed better and had more trust in three-state alarms compared to two-state alarms. Furthermore, these findings have significant theoretical implications and practical applications as they apply to improving the efficiency and effectiveness of nuclear power plant operations.

  1. Foragers of sympatric Asian honey bee species intercept competitor signals by avoiding benzyl acetate from Apis cerana alarm pheromone.

    Science.gov (United States)

    Wen, Ping; Cheng, Yanan; Qu, Yufeng; Zhang, Hongxia; Li, Jianjun; Bell, Heather; Tan, Ken; Nieh, James

    2017-07-27

    While foraging, animals can form inter- and intraspecific social signalling networks to avoid similar predators. We report here that foragers of different native Asian honey bee species can detect and use a specialized alarm pheromone component, benzyl acetate (BA), to avoid danger. We analysed the volatile alarm pheromone produced by attacked workers of the most abundant native Asian honey bee, Apis cerana and tested the responses of other bee species to these alarm signals. As compared to nest guards, A. cerana foragers produced 3.38 fold higher levels of BA. In foragers, BA and (E)-dec-2-en-1-yl acetate (DA) generated the strongest antennal electrophysiological responses. BA was also the only compound that alerted flying foragers and inhibited A. cerana foraging. BA thereby decreased A. cerana foraging for risky sites. Interestingly, although BA occurs only in trace amounts and is nearly absent in sympatric honeybee species (respectively only 0.07% and 0.44% as much in A. dorsata and A. florea), these floral generalists detected and avoided BA as strongly as they did to their own alarm pheromone on natural inflorescences. These results demonstrate that competing pollinators can take advantage of alarm signal information provided by other species.

  2. Interevent times in a new alarm-based earthquake forecasting model

    Science.gov (United States)

    Talbi, Abdelhak; Nanjo, Kazuyoshi; Zhuang, Jiancang; Satake, Kenji; Hamdache, Mohamed

    2013-09-01

    This study introduces a new earthquake forecasting model that uses the moment ratio (MR) of the first to second order moments of earthquake interevent times as a precursory alarm index to forecast large earthquake events. This MR model is based on the idea that the MR is associated with anomalous long-term changes in background seismicity prior to large earthquake events. In a given region, the MR statistic is defined as the inverse of the index of dispersion or Fano factor, with MR values (or scores) providing a biased estimate of the relative regional frequency of background events, here termed the background fraction. To test the forecasting performance of this proposed MR model, a composite Japan-wide earthquake catalogue for the years between 679 and 2012 was compiled using the Japan Meteorological Agency catalogue for the period between 1923 and 2012, and the Utsu historical seismicity records between 679 and 1922. MR values were estimated by sampling interevent times from events with magnitude M ≥ 6 using an earthquake random sampling (ERS) algorithm developed during previous research. Three retrospective tests of M ≥ 7 target earthquakes were undertaken to evaluate the long-, intermediate- and short-term performance of MR forecasting, using mainly Molchan diagrams and optimal spatial maps obtained by minimizing forecasting error defined by miss and alarm rate addition. This testing indicates that the MR forecasting technique performs well at long-, intermediate- and short-term. The MR maps produced during long-term testing indicate significant alarm levels before 15 of the 18 shallow earthquakes within the testing region during the past two decades, with an alarm region covering about 20 per cent (alarm rate) of the testing region. The number of shallow events missed by forecasting was reduced by about 60 per cent after using the MR method instead of the relative intensity (RI) forecasting method. At short term, our model succeeded in forecasting the

  3. Complex programmable logic device based alarm sequencer for nuclear power plants

    International Nuclear Information System (INIS)

    Khedkar, Ravindra; Solomon, J. Selva; KrishnaKumar, B.

    2001-01-01

    Complex Programmable Logic Device based Alarm Sequencer is an instrument, which detects alarms, memorizes them and displays the sequences of occurrence of alarms. It caters to sixteen alarm signals and distinguishes the sequence among any two alarms with a time resolution of 1 ms. The system described has been designed for continuous operation in process plants, nuclear power plants etc. The system has been tested and found to be working satisfactorily. (author)

  4. Design criteria and principles for criticality detection and alarm systems

    International Nuclear Information System (INIS)

    Delafield, H.J.; Clifton, J.J.

    1984-10-01

    The report gives design principles and criteria for criticality detection and alarm systems based on earlier work and revised in the light of more recent experience. In particular, account is taken of the developments which have taken place in the field of radiation detection and in the understanding of the different types of criticality excursion. General guidance is given on the principles to apply in deciding upon the need for a criticality system. The characteristics of a criticality incident are described in terms of the minimum incident of concern, and the radiation field. Criteria for the threshold of detection of a criticality incident are then derived and the methods of detection considered. The selection and siting of criticality detectors is discussed, and design principles are given for alarm systems. Finally, testing and post-alarm procedures are outlined, followed by a summary of the principal recommendations. The supporting Appendices include a discussion of reliability and a summary of radiation detector characteristics. (author)

  5. General multiplex centralized fire-alarm display system

    International Nuclear Information System (INIS)

    Zhu Liqun; Chen Jinming

    2002-01-01

    The fire-alarm display system is developed, which can connect with each type of fire controllers produced in the factory and SIGMASYS controllers. It can display whole alarm information. The display system software gathers communication, database and multimedia, has functions of inspecting fire, showing alarm, storing data, searching information and so on. The drawing software lets the user expediently add, delete, move and modify fire detection or fire fighting facilities on the building floor maps. The graphic transform software lets the display use the vectorgraph produced by popular plotting software such as Auto CAD. The system software provides the administration function, such as log book of changing shift and managing workers etc.. The software executed on Windows 98 platform. The user interface is friendly and reliable in operation

  6. Development of a criticality monitoring and alarm system

    International Nuclear Information System (INIS)

    Egey, Julio; Izraelevitch, Federico H.; Matatagui, Emilio

    2009-01-01

    In this work we are presenting the development of a Criticality Monitor and Alarm System (SIMAC). It monitors the burst of radiation produced during such an accident and triggers an alarm for evacuation in case the radiation exceeds a pre-established threshold. It consists of two subsystems, one for gamma rays and the other for neutrons. Each subsystem has three independent detectors modules. Each module is composed of an ion chamber plus its associated electronics, feeding a logic module that in turn would trigger the evacuation alarm. An additional feature is a PC interface for data acquisition. The radiation detectors are ion chambers working in current mode. The electronics associated to each detector can manage a wide signal range using a logarithmic converter. (author)

  7. Level of flexibility through sit and reach test from research performed in São Paulo city doi: 10.5007/1980-0037.2010v12n6p415

    Directory of Open Access Journals (Sweden)

    Cibele Calvi Anic Ribeiro

    2010-09-01

    Full Text Available The objective of this study was to examine the level of flexibility in men and women of different ages by the sit-and-reach test and to classify them according to the Canadian Standardized Test of Fitness (CSTF. The results were used to elaborate a new table that reflects the population studied. The sample consisted of 16,405 physically active and inactive subjects who were divided according to age: 15 to 19 (n = 954, 20 to 29 (n = 2916, 30 to 39 (n = 2161, 40 to 49 (n = 2333, 50 to 59 (n = 2739, 60 to 69 (n = 3195, and > 70 years (n = 2107. Percentiles were calculated using the original test scores and the 20th, 40th, 60th and 80th percentiles were used as cut-offs for classification as poor, below the average, average, above the average and excellent, respectively. According to the CSTF classification, the age groups from 15 to 39 years were classified as poor, with mean flexibility ranging from 24.805±9.684 to 26.130± 10.111 cm in women and from 21.480±9.905 to 22.848±9.648 cm in men. In the 40- to 69-year age groups, mean flexibility ranged from 22.768±9.627 to 25.396±9.547 in women and from 16.396±10.136 to 19.935±9.192 cm in men and was classified as below the average. Although most of the subjects performed regular exercise, the mean flexibility level observed did not correspond to the average suggested by the CSTF, demonstrating the importance of elaborating national reference tables and of establishing new normative values such as the table proposed in this study.

  8. PROLOG language application for alarm system realization in accelerator control

    International Nuclear Information System (INIS)

    Frolov, I.; Vaguine, A.; Abe, I.; Nakahara, K.; Furukawa, K.; Kamikubota, N.

    1994-01-01

    Such PROLOG features as backtracking, matching and recursive data representation are powerful tools for ALARM system realization. Although the main idea is the possibility to describe some technical system in recursive form, backtracking and matching are ideal for processing recursive data structures. This paper represents a technique which would allow PROLOG language application for ALARM system realization using an example of the KEK LINAC magnet system. The technique is based on an object-oriented internal data representation in terms of objects, properties, relations and knowledge conception. In addition, each property value is characterized by a typical 'time life'. (author)

  9. AN AUTOMATIC CAR ANTI-THEFT ALARM SYSTEM

    OpenAIRE

    J. A. Enokela; E.J. Ibanga

    2007-01-01

    The theft of cars and other automobiles by criminals has become so frequent in our society as to be classified as alarming. Most of the thefts are organized by gangs of robbers but sometimes individuals engage in this activity. The result usually, however, is that the persons from whom the vehicles have been stolen are left to grieve as many of these vehicles are never recovered. This paper describes a simple alarm system that can be easily installed in all kinds of vehicles. The system desc...

  10. Understanding heart rate alarm adjustment in the intensive care units through an analytical approach.

    Directory of Open Access Journals (Sweden)

    Richard L Fidler

    Full Text Available Heart rate (HR alarms are prevalent in ICU, and these parameters are configurable. Not much is known about nursing behavior associated with tailoring HR alarm parameters to individual patients to reduce clinical alarm fatigue.To understand the relationship between heart rate (HR alarms and adjustments to reduce unnecessary heart rate alarms.Retrospective, quantitative analysis of an adjudicated database using analytical approaches to understand behaviors surrounding parameter HR alarm adjustments. Patients were sampled from five adult ICUs (77 beds over one month at a quaternary care university medical center. A total of 337 of 461 ICU patients had HR alarms with 53.7% male, mean age 60.3 years, and 39% non-Caucasian. Default HR alarm parameters were 50 and 130 beats per minute (bpm. The occurrence of each alarm, vital signs, and physiologic waveforms was stored in a relational database (SQL server.There were 23,624 HR alarms for analysis, with 65.4% exceeding the upper heart rate limit. Only 51% of patients with HR alarms had parameters adjusted, with a median upper limit change of +5 bpm and -1 bpm lower limit. The median time to first HR parameter adjustment was 17.9 hours, without reduction in alarms occurrence (p = 0.57.HR alarms are prevalent in ICU, and half of HR alarm settings remain at default. There is a long delay between HR alarms and parameters changes, with insufficient changes to decrease HR alarms. Increasing frequency of HR alarms shortens the time to first adjustment. Best practice guidelines for HR alarm limits are needed to reduce alarm fatigue and improve monitoring precision.

  11. Alarm Fatigue vs User Expectations Regarding Context-Aware Alarm Handling in Hospital Environments Using CallMeSmart.

    Science.gov (United States)

    Solvoll, Terje; Arntsen, Harald; Hartvigsen, Gunnar

    2017-01-01

    Surveys and research show that mobile communication systems in hospital settings are old and cause frequent interruptions. In the quest to remedy this, an Android based communication system called CallMeSmart tries to encapsulate most of the frequent communication into one hand held device focusing on reducing interruptions and at the same time make the workday easier for healthcare workers. The objective of CallMeSmart is to use context-awareness techniques to automatically monitor the availability of physicians' and nurses', and use this information to prevent or route phone calls, text messages, pages and alarms that would otherwise compromise patient care. In this paper, we present the results from interviewing nurses on alarm fatigue and their expectations regarding context-aware alarm handling using CallMeSmart.

  12. Design of alarm systems in Swedish nuclear power plants

    International Nuclear Information System (INIS)

    Thunberg, Anna; Osvalder, Anna-Lisa

    2008-04-01

    Research within the area of improving alarm system design and performance has mainly focused on new alarm systems. However, smaller modernisations of legacy systems are more common in the Swedish nuclear industry than design of totally new systems. This imposes problems when the new system should function together with the old system. This project deals with the special concerns raised by modernisation projects. The objective of the project has been to increase the understanding of the relationship between the operator's performance and the design of the alarm system. Of major concern has been to consider the cognitive abilities of the operator, different operator roles and work situations, and varying need of information. The aim of the project has been to complement existing alarm design guidance and to develop user-centred alarm design concepts. Different case studies have been performed in several industry sectors (nuclear, oil refining, pulp and paper, aviation and medical care) to identify best practice. Several empirical studies have been performed within the nuclear area to investigate the operator's need of information, performance and workload in different operating modes. The aspect of teamwork has also been considered. The analyses show that the operator has different roles in different work situations which affect both the type of information needed and how the information is processed. In full power operation, the interaction between the operator and the alarm system is driven by internal factors and the operator tries to maintain high situation awareness by actively searching for information. The operator wants to optimise the process and need detailed information with possibilities to follow-up and get historical data. In disturbance management, the operator is more dependent on external information presented by the alarm system. The new compilation of alarm guidance is based on the operator's varying needs in different working situations and is

  13. Automatic duress alarms through physiological response monitoring

    International Nuclear Information System (INIS)

    Roehrig, S.C.

    1977-07-01

    Physiological response monitoring under controlled conditions can provide an effective means for passively determining if the wearer is under moderate to severe stresses. By monitoring the heart rate (HR) and galvanic skin response (GSR) of an individual, it is possible to detect in real time the increase in heart rate and GSR levels due to physiological reactions to mental duress. With existing physiological monitoring equipment, however, the work load of the wearer must be well defined since it is impossible, without additional data, to distinguish mental duress responses from those resulting from moderate physical exertion. Similarly, environmental conditions should be constrained within set limits to avoid masking increases in GSR levels due to metntal stress from those associated with increased perspiration. These constraints should not prove overly restrictive and would allow an integrated security system utilizing physiological monitoring equipment to provide an effective real time, automated early warning system for detection of mental duress or death of the wearer

  14. Metasurface holograms reaching 80% efficiency.

    Science.gov (United States)

    Zheng, Guoxing; Mühlenbernd, Holger; Kenney, Mitchell; Li, Guixin; Zentgraf, Thomas; Zhang, Shuang

    2015-04-01

    Surfaces covered by ultrathin plasmonic structures--so-called metasurfaces--have recently been shown to be capable of completely controlling the phase of light, representing a new paradigm for the design of innovative optical elements such as ultrathin flat lenses, directional couplers for surface plasmon polaritons and wave plate vortex beam generation. Among the various types of metasurfaces, geometric metasurfaces, which consist of an array of plasmonic nanorods with spatially varying orientations, have shown superior phase control due to the geometric nature of their phase profile. Metasurfaces have recently been used to make computer-generated holograms, but the hologram efficiency remained too low at visible wavelengths for practical purposes. Here, we report the design and realization of a geometric metasurface hologram reaching diffraction efficiencies of 80% at 825 nm and a broad bandwidth between 630 nm and 1,050 nm. The 16-level-phase computer-generated hologram demonstrated here combines the advantages of a geometric metasurface for the superior control of the phase profile and of reflectarrays for achieving high polarization conversion efficiency. Specifically, the design of the hologram integrates a ground metal plane with a geometric metasurface that enhances the conversion efficiency between the two circular polarization states, leading to high diffraction efficiency without complicating the fabrication process. Because of these advantages, our strategy could be viable for various practical holographic applications.

  15. Development experience and strategy for the combined algorithm on the alarm processing and diagnosis

    International Nuclear Information System (INIS)

    Chung, Hak-Yeong

    1997-01-01

    In this paper, I presented the development experience on the alarm processing and fault diagnosis which has been achieved from early 1988 to late 1995. The scope covered is the prototype stage, the development stage of on-line operator-aid system, and an intelligent human-machine interface system. In the second part, I proposed a new method (APEXS) of multi-alarm processing to select the causal alarm(s) among occurred alarms by using the time information of each occurred alarm and alarm tree knowledge and the corresponding diagnosis method based on the selected causal alarm(s) by using the prescribed qualitative model. With more knowledge base about the plant and some modification suitable for real environment, APEXS will be able to adapt to a real steam power plant. (author). 18 refs, 3 figs, 1 tab

  16. Integrated alarm annunciation and entry control systems -- Survey results

    International Nuclear Information System (INIS)

    Clever, J.J.; Arakaki, L.H.; Monaco, F.M.; Juarros, L.E.; Quintana, G.R.

    1993-10-01

    This report provides the results and analyses of a detailed survey undertaken in Summer 1993 to address integrated intrusion detection alarm annunciation and entry control system issues. This survey was undertaken as a first attempt toward beginning to answer questions about integrated systems and commercial capabilities to meet or partially meet US Department of Energy (DOE) site needs

  17. Survey of approaches for handling static analysis alarms

    NARCIS (Netherlands)

    Muske, T.; Serebrenik, A.

    2016-01-01

    Static analysis tools have showcased their importance and usefulness in automated detection of code anomalies and defects. However, the large number of alarms reported and cost incurred in their manual inspections have been the major concerns with the usage of static analysis tools. Existing studies

  18. 29 CFR 1910.165 - Employee alarm systems.

    Science.gov (United States)

    2010-07-01

    ... OCCUPATIONAL SAFETY AND HEALTH STANDARDS Fire Protection Other Fire Protection Systems § 1910.165 Employee... notification to assigned personnel whenever a deficiency exists in the system. The employer shall assure that... 29 Labor 5 2010-07-01 2010-07-01 false Employee alarm systems. 1910.165 Section 1910.165 Labor...

  19. Chemical alarm in the termite Termitogeton planus (Rhinotermitidae)

    Czech Academy of Sciences Publication Activity Database

    Dolejšová, Klára; Krasulová, Jana; Kutalová, K.; Hanus, Robert

    2014-01-01

    Roč. 40, 11/12 (2014), s. 1269-1276 ISSN 0098-0331 R&D Projects: GA ČR GAP506/10/1570 Institutional support: RVO:61388963 Keywords : termites * soldiers * frontal gland * alarm pheromone * Rhinotermitidae * Termitogeton Subject RIV: CC - Organic Chemistry Impact factor: 2.747, year: 2014

  20. Communication interface of computerized automatic fire alarm system

    International Nuclear Information System (INIS)

    Yu Hongmei; Zhu Liqun; Fang Shaohong; Du Chengbao

    1997-01-01

    The problems of communication between multiple single-chip computers and microcomputer have been solved by the way of hardware and software. The automatic fire alarm system is realized by using the serial port both on single-chip computer and microcomputer

  1. No more moody mornings : Alarm clock anticipates sleepers' emotions

    NARCIS (Netherlands)

    Wensveen, S.; Overbeeke, K.; Van Kasteren, J.

    2002-01-01

    More eloquent alternatives to the harsh tones of the oldfashioned alarm-clock bells abound, including a newsreaders voice summing up last nights disasters, or a tape of your favourite early morning music. Still, getting out of bed has its difficult moments. All this could well change in the near

  2. Alarm management in TRANSPETRO National Oil Control Center

    Energy Technology Data Exchange (ETDEWEB)

    Amado, Helio; Costa, Luciano [TRANSPETRO - PETROBRAS Transporte S.A., Rio de Janeiro, RJ (Brazil)

    2009-07-01

    For sure Alarm Management is not a new issue. EEMUA 191 has been around since 1999 and everyone has received visits from consultants in this area. Besides this regulators have requested that operators have a policy for it. However there are few papers showing actual pipeline operator experience in alarm management. In this paper we present the work developed in TRANSPETRO National Oil Control Center since 2006, where we operate 5509 km of crude oil and refined products pipelines. Since the beginning of the centralized operation in 2002, alarm management has been a concern but a systematic approach has been taken since 2006. Initially we will make a brief revision of the literature and show trends for regulations. Then we will show the tools and the approach we have taken. Finally, the further developments we see. The point that we want to discuss is that, it has been very difficult to implement the system in a linear way and we believe that companies that have huge legacy systems, the same probably will occur. Putting in simple words, our main conclusion is: Implementing an Alarm Management policy produces good results however probably sometimes is better not to follow strictly the traditional steps. (author)

  3. Automatic diagnosis of multiple alarms for reactor-control rooms

    International Nuclear Information System (INIS)

    Gimmy, K.L.; Nomm, E.

    1981-01-01

    A system has been developed at the Savannah River Plant to help reactor operators respond to multiple alarms in a developing incident situation. The need for such systems has become evident in recent years, particularly after the three Mile Island incident

  4. Satellite Search and Rescue System Studies: Alarm and Position Reporting.

    Science.gov (United States)

    alarm and position reporting ( ALPR ) techniques, and the operational or planned spacecraft which might be available for piggybacking the ALPR payload...Several system concepts were then developed to perform the ALPR functions. The candidates were screened and the preferred system concepts were chosen

  5. SeaQuest/E906 Shift Alarm System

    Science.gov (United States)

    Kitts, Noah

    2014-09-01

    SeaQuest, Fermilab E906, is a fixed target experiment that measures the Drell-Yan cross-section ratio of proton-proton to proton-deuterium collisions in order to extract the sea anti-quark structure of the proton. SeaQuest will extend the measurements made by E866/NuSea with greater precision at higher Bjorken-x. The continuously running experiment is always being monitored. Those on shift must keep track of all of the detector readouts in order to make sure the experiment is running correctly. As an experiment that is still in its early stages of running, an alarm system for people on shift is being created to provide warnings, such as a plot showing a detector's performance is sufficiently different to need attention. This plan involves python scripts that track live data. When the data shows a problem within the experiment, a corresponding alarm ID is sent to the MySQL database which then sets off an alarm. These alarms, which will alert the person on shift through both an audible and visual response, are important for ensuring that issues do not go unnoticed, and to help make sure the experiment is recording good data.

  6. Wild birds learn to eavesdrop on heterospecific alarm calls.

    Science.gov (United States)

    Magrath, Robert D; Haff, Tonya M; McLachlan, Jessica R; Igic, Branislav

    2015-08-03

    Many vertebrates gain critical information about danger by eavesdropping on other species' alarm calls [1], providing an excellent context in which to study information flow among species in animal communities [2-4]. A fundamental but unresolved question is how individuals recognize other species' alarm calls. Although individuals respond to heterospecific calls that are acoustically similar to their own, alarms vary greatly among species, and eavesdropping probably also requires learning [1]. Surprisingly, however, we lack studies demonstrating such learning. Here, we show experimentally that individual wild superb fairy-wrens, Malurus cyaneus, can learn to recognize previously unfamiliar alarm calls. We trained individuals by broadcasting unfamiliar sounds while simultaneously presenting gliding predatory birds. Fairy-wrens in the experiment originally ignored these sounds, but most fled in response to the sounds after two days' training. The learned response was not due to increased responsiveness in general or to sensitization following repeated exposure and was independent of sound structure. Learning can therefore help explain the taxonomic diversity of eavesdropping and the refining of behavior to suit the local community. In combination with previous work on unfamiliar predator recognition (e.g., [5]), our results imply rapid spread of anti-predator behavior within wild populations and suggest methods for training captive-bred animals before release into the wild [6]. A remaining challenge is to assess the importance and consequences of direct association of unfamiliar sounds with predators, compared with social learning-such as associating unfamiliar sounds with conspecific alarms. Copyright © 2015 Elsevier Ltd. All rights reserved.

  7. Cell broadcast trials in The Netherlands: Using mobile phone technology for citizens' alarming

    International Nuclear Information System (INIS)

    Jagtman, H.M.

    2010-01-01

    In emergency situations authorities need to warn the public. The conventionally used method for warning citizens in The Netherlands is the use of a siren. Modern telecommunication technologies, especially the use of text-based features of mobile phones, have great potential for warning the public. In the years 2005-2007 cell broadcast was tested during several large-scale field trials with citizens in The Netherlands. One of the questions was to determine the penetration of cell broadcast for citizens' alarming. This article argues that the definition of penetration in the light of warning citizens in case of emergencies should include the citizens' responses to warning messages. In addition, the approach to determining the penetration, the data and validity issues regarding these data is discussed. The trials have shown cell broadcast has potential to become an effective citizens' alarming technology. This however requires the entire technological and organisational chain of the warning system to function correctly. Attention is required to network management, handset improvements and correct communication to the public about the conditions under which a cell broadcast message can be received. The latter includes managing realistic expectations including circumstances in which cell broadcast will not reach a citizen.

  8. Prioritizing alarms from sensor-based detection models in livestock production - A review on model performance and alarm reducing methods

    DEFF Research Database (Denmark)

    Dominiak, Katarina Sylow; Kristensen, Anders Ringgaard

    2017-01-01

    The objective of this review is to present, evaluate and discuss methods for reducing false alarms in sensor-based detection models developed for livestock production as described in the scientific literature. Papers included in this review are all peer-reviewed and present sensor-based detection...

  9. [Optimized resource mobilization and quality of treatment of severely injured patients through a structured trauma room alarm system].

    Science.gov (United States)

    Spering, C; Roessler, M; Kurlemann, T; Dresing, K; Stürmer, K M; Lehmann, W; Sehmisch, S

    2017-12-12

    The treatment of severely injured patients in the trauma resuscitation unit (TRU) requires an interdisciplinary and highly professional trauma team approach. The complete team needs to be waiting for the patient in the TRU on arrival. Treating severely injured patients in the TRU, the trauma team connects the initial preclinical emergency stabilization with the subsequent sophisticated treatment. Thus, the trauma team depends on concise information from the emergency personnel at the scene to provide its leader with further information as well as an accurate alarm including all departments necessary to stabilize the patient in the TRU. Aiming at an accurate and most efficient trauma team alarm, this study was designed to provide and analyze an alarm system which mobilizes the trauma team in a stepwise fashion depending on the pattern of injuries and the threat to life. The trauma team alarm system was analyzed in a prospective data acquisition at a level I trauma center over a period of 12 months. Evaluation followed the acquisition phase and provided comparison to the status prior to the establishment of the alarm system. All items underwent statistical testing using t‑tests (p  0.01). The duration from arrival of the patient in the TRU to the initial multislice computed tomography (CT) scan was reduced by 6 min while the total period of treatment in the TRU was reduced by 17 min. After the alarm system to gradually mobilize the trauma team was put into action, team members left the TRU if unneeded prior to finishing the initial treatment in only 4% of the cases. The patient fatality rate was 8.8% (injury severity score, ISS = 23 points) after establishment of the alarm system compared to 12.9% (ISS = 25 points) before. The implementation of an accurate and patient status-based alarm system to mobilize the trauma team can improve the quality of treatment while the duration of treatment of the severely injured patients in the TRU can be decreased. It

  10. Reaching ignition in the tokamak

    International Nuclear Information System (INIS)

    Furth, H.P.

    1985-06-01

    This review covers the following areas: (1) the physics of burning plasmas, (2) plasma physics requirements for reaching ignition, (3) design studies for ignition devices, and (4) prospects for an ignition project

  11. Retrofitting alarm prioritization at Bruce A: strategy development and implementation experience

    International Nuclear Information System (INIS)

    Davey, E.; Hickey, D.; Babcock, B.

    1997-01-01

    A prioritization strategy for computer-displayed control room alarms has been developed for Bruce A to better assist operations staff in visually identifying key alarms and judging the relative importance of alarms. The strategy consists of assigning each alarm indicative of a problem to be addressed to one of five priority categories. Each alarm is assigned to an alarm category based on an off-line analysis of the consequence and response characteristics applicable to the alarm for three plant operating contexts. The colour of the alarm message is used to convey the priority category of each alarm in computer-based alarm displays. In addition, alarms indicative of non-problematic changes in the state of plant equipment and processes are given a separate colour assignment to visually differentiate them from alarms indicative of problems. This paper outlines the user-based approach employed in the prioritization strategy development, describes the key features of the prioritization strategy adopted, and discusses the initial experience in systematically determining the priority assignments for all 6000 computer-based alarms associated with each generating unit. (author)

  12. The Importance of Technical Reachback in the Adjudication of Radiation Alarms

    International Nuclear Information System (INIS)

    Buckley, W.M.; Allen, R.W.

    2009-01-01

    The large-scale deployment of radiation sensors at borders, ports-of-entry and other locations carries two disparate priorities: the reliable detection and identification of threat materials and the rapid characterization of non-threat materials comprised of naturally occurring radioactive materials (NORM) and legitimate radioactive materials in streams of commerce. These priorities are partially achieved through the technologies contained in the detection systems and the procedures developed for their operation. However, questions and ambiguities will occur. Without established capabilities and procedures for the operators of these detector systems to 'reach back' to trained spectroscopists and appropriate subject matter experts, the system will likely experience an unacceptable number of response operations and delays resolving alarms. Technical reachback operations need to be able to address the priorities discussed above while causing minimal perturbations in the flow of legitimate streams of commerce. Yet when necessary, reachback needs to be able to rapidly mobilize the appropriate response assets.

  13. Stop the Noise: A Quality Improvement Project to Decrease Electrocardiographic Nuisance Alarms.

    Science.gov (United States)

    Sendelbach, Sue; Wahl, Sharon; Anthony, Anita; Shotts, Pam

    2015-08-01

    As many as 99% of alarm signals may not need any intervention and can result in patients' deaths. Alarm management is now a Joint Commission National Patient Safety Goal. To reduce the number of nuisance electrocardiographic alarm signals in adult patients on the medical cardiovascular care unit. A quality improvement process was used that included eliminating duplicative alarms, customizing alarms, changing electrocardiography electrodes daily, standardizing skin preparation, and using disposable electrocardiography leads. In the cardiovascular care unit, the mean number of electrocardiographic alarm signals per day decreased from 28.5 (baseline) to 3.29, an 88.5% reduction. Use of a bundled approach to managing alarm signals decreased the mean number of alarm signals in a cardiovascular care unit. ©2015 American Association of Critical-Care Nurses.

  14. Reducing alarms and prioritising interventions in pig production by simultaneous monitoring of water consumption in multiple pens

    DEFF Research Database (Denmark)

    Dominiak, Katarina Nielsen; Hindsborg, Jeff; Pedersen, L. J.

    to be generated at pen level or merged at section or herd level to reduce the number of alarms. Information on which specific pens or sections are of higher risk of stress or diseases is communicated to the farmer and target work effort to pens at risk. For Herd A, all model parameters defined at section level...... resulted in the best fit (MSE =13.85 litres2/hour). For Herd B, parameters defined at both pen and section level resulted in the best fit (MSE = 1.47 litres2/hour). For both Herd A and Herd B, preliminary results support the spatial approach by generating a reduced number of alarms when comparing section...

  15. Teaching methodology of the diagnosing process on the example of the fire alarm system

    Directory of Open Access Journals (Sweden)

    Paś Jacek

    2017-03-01

    Full Text Available The article presents a method of teaching the process of diagnosing the technical and functional condition of the fire alarm system (SSP. The fire alarm system’s laboratory model is a representation of a real fire alarm system. The lecturer has the opportunity to inflict several different independent damage. The aim of the laboratory exercise is to familiarize students with the methodology and structure of the fire alarm system diagnosing process.

  16. PT-SAFE: a software tool for development and annunciation of medical audible alarms.

    Science.gov (United States)

    Bennett, Christopher L; McNeer, Richard R

    2012-03-01

    Recent reports by The Joint Commission as well as the Anesthesia Patient Safety Foundation have indicated that medical audible alarm effectiveness needs to be improved. Several recent studies have explored various approaches to improving the audible alarms, motivating the authors to develop real-time software capable of comparing such alarms. We sought to devise software that would allow for the development of a variety of audible alarm designs that could also integrate into existing operating room equipment configurations. The software is meant to be used as a tool for alarm researchers to quickly evaluate novel alarm designs. A software tool was developed for the purpose of creating and annunciating audible alarms. The alarms consisted of annunciators that were mapped to vital sign data received from a patient monitor. An object-oriented approach to software design was used to create a tool that is flexible and modular at run-time, can annunciate wave-files from disk, and can be programmed with MATLAB by the user to create custom alarm algorithms. The software was tested in a simulated operating room to measure technical performance and to validate the time-to-annunciation against existing equipment alarms. The software tool showed efficacy in a simulated operating room environment by providing alarm annunciation in response to physiologic and ventilator signals generated by a human patient simulator, on average 6.2 seconds faster than existing equipment alarms. Performance analysis showed that the software was capable of supporting up to 15 audible alarms on a mid-grade laptop computer before audio dropouts occurred. These results suggest that this software tool provides a foundation for rapidly staging multiple audible alarm sets from the laboratory to a simulation environment for the purpose of evaluating novel alarm designs, thus producing valuable findings for medical audible alarm standardization.

  17. Potential Nematode Alarm Pheromone Induces Acute Avoidance in Caenorhabditis elegans.

    Science.gov (United States)

    Zhou, Ying; Loeza-Cabrera, Mario; Liu, Zheng; Aleman-Meza, Boanerges; Nguyen, Julie K; Jung, Sang-Kyu; Choi, Yuna; Shou, Qingyao; Butcher, Rebecca A; Zhong, Weiwei

    2017-07-01

    It is crucial for animal survival to detect dangers such as predators. A good indicator of dangers is injury of conspecifics. Here we show that fluids released from injured conspecifics invoke acute avoidance in both free-living and parasitic nematodes. Caenorhabditis elegans avoids extracts from closely related nematode species but not fruit fly larvae. The worm extracts have no impact on animal lifespan, suggesting that the worm extract may function as an alarm instead of inflicting physical harm. Avoidance of the worm extract requires the function of a cGMP signaling pathway that includes the cGMP-gated channel TAX-2/TAX-4 in the amphid sensory neurons ASI and ASK. Genetic evidence indicates that the avoidance behavior is modulated by the neurotransmitters GABA and serotonin, two common targets of anxiolytic drugs. Together, these data support a model that nematodes use a nematode-specific alarm pheromone to detect conspecific injury. Copyright © 2017 by the Genetics Society of America.

  18. Plant experience with an expert system for alarm diagnosis

    International Nuclear Information System (INIS)

    Gimmy, K.L.

    1986-01-01

    An expert system called Diagnosis of Multiple Alarms (DMA) is in routine use at four nuclear reactors operated by the DuPont Company. The system is wired to plant alarm annunciators and does event-tree analysis to see if a pattern exists. Any diagnosis is displayed to the plant operator and the corrective procedure to be followed is also identified. The display is automatically superseded if a higher priority diagnosis is made. The system is integrated with operator training and procedures. Operating results have been positive. DMA has diagnosed several hard-to-locate small leaks. There have been some false diagnosis, and realistic plant environments must be considered in such expert systems. 2 refs., 5 figs

  19. Fault tolerant microcomputer based alarm annunciator for Dhruva reactor

    International Nuclear Information System (INIS)

    Chandra, A.K.

    1988-01-01

    The Dhruva alarm annunciator displays the status of 624 alarm points on an array of display windows using the standard ringback sequence. Recognizing the need for a very high availability, the system is implemented as a fault tolerant configuration. The annunciator is partitioned into three identical units; each unit is implemented using two microcomputers wired in a hot standby mode. In the event of one computer malfunctioning, the standby computer takes over control in a bouncefree transfer. The use of microprocessors has helped built-in flexibility in the system. The system also provides built-in capability to resolve the sequence of occurrence of events and conveys this information to another system for display on a CRT. This report describes the system features, fault tolerant organisation used and the hardware and software developed for the annunciation function. (author). 8 figs

  20. Central display system of figures in fire alarm

    International Nuclear Information System (INIS)

    Fang Shaohong; Zhu Zicheng; Zhu Liqun; Ren Yi; Yu Hongmei; Du Chengbao; Xie Guoxue

    1997-01-01

    A new type of 'central display system of figures in fire alarm' includes two parts: (1) software package of drawing picture; (2) real time processing and operate system (POS). Main function of the software package is to draw floor plane figures, fire-fighting facility signs and room numbers; and then all pictures are used in POS. Main functions of POS are to process fire alarm, faults and activation of fire fighting control facility, save and print reports, look over floor plane figures, look over concrete condition of fire fighting facilities, and to show appropriate prompt according to different case. This system realizes many functions, such as, control with mouse, operation with push-button, menu operation interface, flip windows to prompt, and chinese character. It have won acclaim for its amazing interface, its convenience to operate, its reliability and flexibility

  1. Design of reactor alarm instrument based on SOPC

    International Nuclear Information System (INIS)

    Li Meng; Lu Yi; Rong Ru

    2008-01-01

    The design of embedded alarm instrument in reactors based on Nios II CPU is introduced in this paper. This design uses the SOPC technology based on the Cyclone series FPGA as a digital bench, and connects the MPU and drivers and interface of times, RS232, sdram,and etc. into a FPGA chip. It is proved that the system achieves the design goals in primary experimentation. (authors)

  2. Process plant alarm diagnosis using synthesised fault tree knowledge

    International Nuclear Information System (INIS)

    Trenchard, A.J.

    1990-01-01

    The development of computer based tools, to assist process plant operators in their task of fault/alarm diagnosis, has received much attention over the last twenty five years. More recently, with the emergence of Artificial Intelligence (AI) technology, the research activity in this subject area has heightened. As a result, there are a great variety of fault diagnosis methodologies, using many different approaches to represent the fault propagation behaviour of process plant. These range in complexity from steady state quantitative models to more abstract definitions of the relationships between process alarms. Unfortunately, very few of the techniques have been tried and tested on process plant and even fewer have been judged to be commercial successes. One of the outstanding problems still remains the time and effort required to understand and model the fault propagation behaviour of each considered process. This thesis describes the development of an experimental knowledge based system (KBS) to diagnose process plant faults, as indicated by process variable alarms. In an attempt to minimise the modelling effort, the KBS has been designed to infer diagnoses using a fault tree representation of the process behaviour, generated using an existing fault tree synthesis package (FAULTFINDER). The process is described to FAULTFINDER as a configuration of unit models, derived from a standard model library or by tailoring existing models. The resultant alarm diagnosis methodology appears to work well for hard (non-rectifying) faults, but is likely to be less robust when attempting to diagnose intermittent faults and transient behaviour. The synthesised fault trees were found to contain the bulk of the information required for the diagnostic task, however, this needed to be augmented with extra information in certain circumstances. (author)

  3. The research of nuclear experiment radiation environment wireless alarm device

    International Nuclear Information System (INIS)

    Wang Xiaoqiong; Wang Pan; Fang Fang

    2009-01-01

    This article introduces based on monolithic integrated circuit's nuclear experiment radiation environment wireless alarm device's software and hardware design. The system by G-M tube, high-pressured module, signal conditioning circuit, power source module, monolithic integrated circuit and wireless transmission module is composed. The device has low power consumption, high performance, high accuracy detection, easy maintenance, small size, simple operation, and other features, and has a broad application prospects. (authors)

  4. Functional relationship based alarm processing system for nuclear power

    International Nuclear Information System (INIS)

    Corsberg, D.; Sebo, D.

    1986-01-01

    At the Idaho National Engineering Laboratory (INEL), two knowledge-based systems, Alarm Filtering System (AFS) and Reactor Safety Assessment System (RSAS), are being developed to address the problem of information overload faced by operators during major plant transients. This paper discusses those systems, their strengths and weaknesses, and how their basic methodologies can be combined into a system that can better support the operator in identifying the plant state

  5. Recent and proposed changes in criticality alarm system requirements

    International Nuclear Information System (INIS)

    Putman, V.L.

    1998-01-01

    Various changes in criticality alarm system (CAS) requirements of American Nuclear Society (ANS) standards, US Department of Energy (DOE) orders, US Nuclear Regulatory Commission (NRC) regulations and guidance, and Occupational Safety and Health Administration (OSHA) standards or regulations were approved or proposed in the last 5 yr. Many changes interpreted or clarified existing requirements or accommodated technological or organizational developments. However, some changes could substantively affect CAS programs, including several changes originally thought to be editorial. These changes are discussed here

  6. Evaluation of personal alarm devices for fire fighters

    International Nuclear Information System (INIS)

    Sharry, J.; da Roza, R.A.

    1984-01-01

    Although three of the models of the personal alarm devices (PAD) tested had received letters of acceptability and compliance from the State of California/Occupational Safety and Health Administration (Cal/OSHA), it was found that none of the models met all of the Cal/OSHA specifications. Additional tests showed other deficiencies in the PADs' design. This points out the need for the purchaser or user of such devices to specify acceptance criteria and perform his own tests

  7. Cancer prevention and control: alarming challenges in China

    OpenAIRE

    Bode, Ann M.; Dong, Zigang; Wang, Hongyang

    2015-01-01

    China is geographically the third largest country in the world and the most populated low-to-middle-income country. Cancer incidence and mortality rates for some cancers in the USA and European countries have steadily decreased over the last decades, whereas the incidence and mortality of certain cancers in China have been increasing at an alarming speed. Rapid industrialization and urbanization in China have been accompanied by incredible changes in lifestyle and environment combined with an...

  8. Reduction of false arrhythmia alarms using signal selection and machine learning

    NARCIS (Netherlands)

    Eerikäinen, L.M.; Vanschoren, J.; Rooijakkers, M.J.; Vullings, R.; Aarts, R.M.

    2016-01-01

    In this paper, we propose an algorithm that classifies whether a generated cardiac arrhythmia alarm is true or false. The large number of false alarms in intensive care is a severe issue. The noise peaks caused by alarms can be high and in a noisy environment nurses can experience stress and

  9. Project 93L-EWL-097, fire alarm system improvements, 300 Area

    International Nuclear Information System (INIS)

    Scott, M.V.

    1995-01-01

    This document contains the Acceptance Test Procedure (ATP) which will demonstrate that the modifications to the Fire Protection systems in the 338 Building function as intended. The ATP will test the fire alarm control panel, flow alarm pressure switch, post indicator valve tamper switch, heat detectors, flow switches, and fire alarm signaling devices

  10. Spiny lobsters detect conspecific blood-borne alarm cues exclusively through olfactory sensilla.

    Science.gov (United States)

    Shabani, Shkelzen; Kamio, Michiya; Derby, Charles D

    2008-08-01

    When attacked by predators, diverse animals actively or passively release molecules that evoke alarm and related anti-predatory behavior by nearby conspecifics. The actively released molecules are alarm pheromones, whereas the passively released molecules are alarm cues. For example, many insects have alarm-signaling systems that involve active release of alarm pheromones from specialized glands and detection of these signals using specific sensors. Many crustaceans passively release alarm cues, but the nature of the cues, sensors and responses is poorly characterized. Here we show in laboratory and field experiments that injured Caribbean spiny lobsters, Panulirus argus, passively release alarm cues via blood (hemolymph) that induce alarm responses in the form of avoidance and suppression of feeding. These cues are detected exclusively through specific olfactory chemosensors, the aesthetasc sensilla. The alarm cues for Caribbean spiny lobsters are not unique to the species but do show some phylogenetic specificity: P. argus responds primarily with alarm behavior to conspecific blood, but with mixed alarm and appetitive behaviors to blood from the congener Panulirus interruptus, or with appetitive behaviors to blood from the blue crab Callinectes sapidus. This study lays the foundation for future neuroethological studies of alarm cue systems in this and other decapod crustaceans.

  11. 47 CFR 80.307 - Compulsory use of radiotelegraph auto alarm.

    Science.gov (United States)

    2010-10-01

    ... 47 Telecommunication 5 2010-10-01 2010-10-01 false Compulsory use of radiotelegraph auto alarm. 80.307 Section 80.307 Telecommunication FEDERAL COMMUNICATIONS COMMISSION (CONTINUED) SAFETY AND SPECIAL... Safety Watches § 80.307 Compulsory use of radiotelegraph auto alarm. The radiotelegraph auto alarm...

  12. Development of the effectiveness measure for an advanced alarm system using signal detection theory

    International Nuclear Information System (INIS)

    Park, J.K.; Choi, S.S.; Hong, J.H.; Chang, S.H.

    1997-01-01

    Since many alarms which are activated during major process deviations or accidents in nuclear power plants can result in negative effects for operators, various types of advanced alarm systems that can select important alarms for the identification of process deviation have been developed to reduce the operator's workload. However, the irrelevant selection of important alarms could distract the operator from correct identification of process deviation. Therefore, to evaluate the effectiveness of the advanced alarm system, a tradeoff between the alarm reduction rate (how many alarms are reduced?) and informativeness (how many important alarms that are conducive to identifying process deviation are provided?) of an advanced alarm system should be considered. In this paper, a new measure is proposed to evaluate the effectiveness of an advanced alarm system with regard to the identification of process deviation. Here, the effectiveness measure is the combination of informativeness measure and reduction rate, and the informativeness measure means the information processing capability performed by the advanced alarm system including wrong rejection and wrong acceptance, and it can be calculated using the signal detection theory (SDT). The effectiveness of the prototype alarm system was evaluated using the loss of coolant accident (LOCA) scenario, and the validity of the effectiveness measure was investigated from two types of the operator response, such as the identification accuracy and the operator's preference for the identification of LOCA

  13. Moving Towards a Common Alarm Service for the LHC Era

    CERN Document Server

    Calderini, F; Stapley, N; Tyrell, M W

    2003-01-01

    The Large Hadron Collider (LHC) is one of the greatest technological challenges ever faced by accelerator builders. It is due for commissioning in 4 years and will have a lifetime well in excess of 10. The LHC will contain a completely heterogeneous mixture of industrial controls, both hardware and software, as well as dedicated, specialised, 'home' built systems. As part of the control infrastructure of such a complex machine, a number of 'services' will be essential as aids during operation, such as: logging / archiving, post-mortem, sequences, alarm system, etc. This paper describes the approach to be taken in order to define and provide the alarm service necessary for LHC. Details will be given of: the graceful transition from the current LEP alarm system; accommodating the SPS, PS and CERN's technical services; the technologies to be used; the approach of parallel investigations of industrial and 'home' builtsystems to ensure the best possible solution; and an indication of time scales to provide an oper...

  14. [Alarm symptoms of meningitis in children with fever].

    Science.gov (United States)

    Geurts, Dorien H F; Moll, Henriette A

    2011-01-01

    A 15-year-old girl presented with fever and pain in her legs. A viral infection was suspected, but within 24 hours she became confused and developed meningeal signs, based on which she was diagnosed as having meningitis. Within a few hours a 6-month-old boy developed fever, a grey colour, bulging fontanel, cold hands and feet, and was groaning. He too appeared to have meningitis. It is important to recognize this serious infection in children with fever, since delay of diagnosis and treatment may result in serious complications. Recognition is difficult because of non-specific symptoms on presentation and a lack of alarm symptoms early in the course of the disease. Alarm symptoms of serious infection in children are cyanosis, rapid breathing, decreased capillary refill, petechial rash, meningeal signs, leg pain and decreased consciousness. If serious infection is uncertain in a child with fever, parents should be advised on the potential course of the disease, the alarm symptoms and the need to seek medical help in time.

  15. Masters Thesis- Criticality Alarm System Design Guide with Accompanying Alarm System Development for the Radioisotope Production Laboratory in Richland, Washington

    Energy Technology Data Exchange (ETDEWEB)

    Greenfield, Bryce A. [Univ. of New Mexico, Albuquerque, NM (United States)

    2009-12-01

    A detailed instructional manual was created to guide criticality safety engineers through the process of designing a criticality alarm system (CAS) for Department of Energy (DOE) hazard class 1 and 2 facilities. Regulatory and technical requirements were both addressed. A list of design tasks and technical subtasks are thoroughly analyzed to provide concise direction for how to complete the analysis. An example of the application of the design methodology, the Criticality Alarm System developed for the Radioisotope Production Laboratory (RPL) of Richland, Washington is also included. The analysis for RPL utilizes the Monte Carlo code MCNP5 for establishing detector coverage in the facility. Significant improvements to the existing CAS were made that increase the reliability, transparency, and coverage of the system.

  16. RECORDS REACHING RECORDING DATA TECHNOLOGIES

    Directory of Open Access Journals (Sweden)

    G. W. L. Gresik

    2013-07-01

    Full Text Available The goal of RECORDS (Reaching Recording Data Technologies is the digital capturing of buildings and cultural heritage objects in hard-to-reach areas and the combination of data. It is achieved by using a modified crane from film industry, which is able to carry different measuring systems. The low-vibration measurement should be guaranteed by a gyroscopic controlled advice that has been , developed for the project. The data were achieved by using digital photography, UV-fluorescence photography, infrared reflectography, infrared thermography and shearography. Also a terrestrial 3D laser scanner and a light stripe topography scanner have been used The combination of the recorded data should ensure a complementary analysis of monuments and buildings.

  17. Records Reaching Recording Data Technologies

    Science.gov (United States)

    Gresik, G. W. L.; Siebe, S.; Drewello, R.

    2013-07-01

    The goal of RECORDS (Reaching Recording Data Technologies) is the digital capturing of buildings and cultural heritage objects in hard-to-reach areas and the combination of data. It is achieved by using a modified crane from film industry, which is able to carry different measuring systems. The low-vibration measurement should be guaranteed by a gyroscopic controlled advice that has been , developed for the project. The data were achieved by using digital photography, UV-fluorescence photography, infrared reflectography, infrared thermography and shearography. Also a terrestrial 3D laser scanner and a light stripe topography scanner have been used The combination of the recorded data should ensure a complementary analysis of monuments and buildings.

  18. Human factors engineering guidance for the review of advanced alarm systems

    International Nuclear Information System (INIS)

    O'Hara, J.M.; Brown, W.S.; Higgins, J.C.; Stubler, W.F.

    1994-09-01

    This report provides guidance to support the review of the human factors aspects of advanced alarm system designs in nuclear power plants. The report is organized into three major sections. The first section describes the methodology and criteria that were used to develop the design review guidelines. Also included is a description of the scope, organization, and format of the guidelines. The second section provides a systematic review procedure in which important characteristics of the alarm system are identified, described, and evaluated. The third section provides the detailed review guidelines. The review guidelines are organized according to important characteristics of the alarm system including: alarm definition; alarm processing and reduction; alarm prioritization and availability; display; control; automated; dynamic, and modifiable characteristics; reliability, test, maintenance, and failure indication; alarm response procedures; and control-display integration and layout

  19. Human factors engineering guidance for the review of advanced alarm systems

    Energy Technology Data Exchange (ETDEWEB)

    O`Hara, J.M.; Brown, W.S.; Higgins, J.C.; Stubler, W.F. [Brookhaven National Lab., Upton, NY (United States)

    1994-09-01

    This report provides guidance to support the review of the human factors aspects of advanced alarm system designs in nuclear power plants. The report is organized into three major sections. The first section describes the methodology and criteria that were used to develop the design review guidelines. Also included is a description of the scope, organization, and format of the guidelines. The second section provides a systematic review procedure in which important characteristics of the alarm system are identified, described, and evaluated. The third section provides the detailed review guidelines. The review guidelines are organized according to important characteristics of the alarm system including: alarm definition; alarm processing and reduction; alarm prioritization and availability; display; control; automated; dynamic, and modifiable characteristics; reliability, test, maintenance, and failure indication; alarm response procedures; and control-display integration and layout.

  20. Intelligent software solution for reliable high efficiency/low false alarm border monitoring

    International Nuclear Information System (INIS)

    Rieck, W.; Iwatschenko, M.

    2001-01-01

    Full text: Radioactivity Monitoring at border stations requires detection systems that are reliably operating under special conditions such as: different types and shapes of vehicles; different velocities; stop and go traffic. ESM has developed a solution that achieves under all such conditions the lowest possible detection limit and avoids false alarms generated by naturally occurring radioactive material (NORM). NBR (Natural Background Reduction) data evaluation - One of the main reasons for the success of the ESM gate monitors is the unique and proprietary NBR-technology of instantaneous discrimination of artificial and natural gamma radiation using large area plastic scintillators. Thus the FHT 1388 gate monitors show 2 unique features: Possible setting of different alarm levels for NORM and artificial gamma sources; Self adjusting compensation of the background shielding of the truck in respect to the detection of artificial sources. Both properties are a preposition for the highly sensitive detection of artificial gamma sources. While at scrap yards and steel mills usually all radioactivity (including NORM) must be detected, the main object of interest in respect to the measuring task at border stations, airports or harbours is clearly the detection of even very small signals of artificial radioactivity. The reliable rejection of the influence of natural radioactivity is of special importance in the case of detection of illicit trafficking, since construction material, fertilisers or soil often lead to much higher detector signals than the alarming levels for dangerous sources of interest. Beside the varying content of natural radioactivity in the load of a truck, different loads and trucks show different influence on the reduction of the ambient radiation due to the passing vehicle. Thus software approaches assuming a specific reduction of the background count rate (regarding relative magnitude and shape) must fail when trucks of different shape and load

  1. Semi-supervised detection of intracranial pressure alarms using waveform dynamics

    International Nuclear Information System (INIS)

    Scalzo, Fabien; Hu, Xiao

    2013-01-01

    Patient monitoring systems in intensive care units (ICU) are usually set to trigger alarms when abnormal values are detected. Alarms are generated by threshold-crossing rules that lead to high false alarm rates. This is a recognized issue that causes alarm fatigue, waste of human resources, and increased patient risks. Recently developed smart alarm models require alarms to be validated by experts during the training phase. The manual annotation process involved is time-consuming and virtually impossible to achieve for the thousands of alarms recorded in the ICU every week. To tackle this problem, we investigate in this study if the use of semi-supervised learning methods, that can naturally integrate unlabeled data samples in the model, can be used to improve the accuracy of the alarm detection. As a proof of concept, the detection system is evaluated on intracranial pressure (ICP) signal alarms. Specific morphological and trending features are extracted from the ICP signal waveform to capture the dynamic of the signal prior to alarms. This study is based on a comprehensive dataset of 4791 manually labeled alarms recorded from 108 neurosurgical patients. A comparative analysis is provided between kernel spectral regression (SR-KDA) and support vector machine (SVM) both modified for the semi-supervised setting. Results obtained during the experimental evaluations indicate that the two models can significantly reduce false alarms using unlabeled samples; especially in the presence of a restrained number of labeled examples. At a true alarm recognition rate of 99%, the false alarm reduction rates improved from 9% (supervised) to 27% (semi-supervised) for SR-KDA, and from 3% (supervised) to 16% (semi-supervised) for SVM. (paper)

  2. Electrochemical nano biosensor alarm devices for the determination of endocrine disruptor agents

    International Nuclear Information System (INIS)

    Iwuoha, E.; Hendricks, N.; Baker, P.

    2009-01-01

    The role of cytochrome P450 (CYP) enzyme systems in the detoxification of bioactive and hydrophobic xenobiotics, such as drugs, environmental pollutants, food supplements, steroids and endocrine disruptors, cannot be over-emphasized. In this study we present the development and amperometric transduction of cytochromal biosensor alarm device for the determination of endocrine disruptors. As a class II microsomal b-type heme enzyme, CYP3A4 requires the obligatory presence of electron transfer donor redox protein, NAD(P)H, and cytochrome b5 for its physiological reactivity. Optimal reconstitution assays preferably involves vesicle forming phospholipids, detergents and specialized reducing agents. Biosensor offers the possibility of observing direct electron transfer reaction of cytochrome P450-3A4 (CYP3A4) without the requirement of the enzyme's physiological redox partners (1,2). In this study, a nanobiosensor alarm device for the determination of 2,4-dichlorophenol (an endocrine disruptor and hepatocarcinogen) was developed with genetically engineered CYP3A4 imprinted on carbon electrode chips that was modified with polypyrrole-gold nanoparticles. The sensor amperometric signals resulted from the two-electron monooxygenation reaction between the ferri-heme CYP3A4 enzyme and the endocrine disruptor compound. The biosensor was interrogated electrochemically for its ability to detect and report the presence of the endocrine disruptor compound in real time. Accordingly, the response time, sensitivity, storage stability, dynamic linear range and detection limits of the device were evaluated. The biosensor alarm device had a detection limit of 43 ng/L for 2,4-dichlorophenol which is lower than the European Union limit of 300 ng/L for pesticide compounds in ground water; as well as the USA Environmental Protection Agency's drinking water equivalent level (DWEL) of 2000 ng/L (3,4). Chromatographic studies despite their tedious sample preparation and time-consuming pre

  3. Evaluation of community-level interventions to increase early initiation of antenatal care in pregnancy: protocol for the Community REACH study, a cluster randomised controlled trial with integrated process and economic evaluations.

    Science.gov (United States)

    Sawtell, Mary; Sweeney, Lorna; Wiggins, Meg; Salisbury, Cathryn; Eldridge, Sandra; Greenberg, Lauren; Hunter, Rachael; Kaur, Inderjeet; McCourt, Christine; Hatherall, Bethan; Findlay, Gail; Morris, Joanne; Reading, Sandra; Renton, Adrian; Adekoya, Ruth; Green, Belinda; Harvey, Belinda; Latham, Sarah; Patel, Kanta; Vanlessen, Logan; Harden, Angela

    2018-03-05

    The provision of high-quality maternity services is a priority for reducing inequalities in health outcomes for mothers and infants. Best practice includes women having their initial antenatal appointment within the first trimester of pregnancy in order to provide screening and support for healthy lifestyles, well-being and self-care in pregnancy. Previous research has identified inequalities in access to antenatal care, yet there is little evidence on interventions to improve early initiation of antenatal care. The Community REACH trial will assess the effectiveness and cost-effectiveness of engaging communities in the co-production and delivery of an intervention that addresses this issue. The study design is a matched cluster randomised controlled trial with integrated process and economic evaluations. The unit of randomisation is electoral ward. The intervention will be delivered in 10 wards; 10 comparator wards will have normal practice. The primary outcome is the proportion of pregnant women attending their antenatal booking appointment by the 12th completed week of pregnancy. This and a number of secondary outcomes will be assessed for cohorts of women (n = approximately 1450 per arm) who give birth 2-7 and 8-13 months after intervention delivery completion in the included wards, using routinely collected maternity data. Eight hospitals commissioned to provide maternity services in six NHS trusts in north and east London and Essex have been recruited to the study. These trusts will provide anonymised routine data for randomisation and outcomes analysis. The process evaluation will examine intervention implementation, acceptability, reach and possible causal pathways. The economic evaluation will use a cost-consequences analysis and decision model to evaluate the intervention. Targeted community engagement in the research process was a priority. Community REACH aims to increase early initiation of antenatal care using an intervention that is co-produced and

  4. Krohne Flow Indicator and High Flow Alarm - Local Indicator and High Flow Alarm of Helium Flow from the SCHe Purge Lines C and D to the Process Vent

    International Nuclear Information System (INIS)

    MISKA, C.R.

    2000-01-01

    Flow Indicators/alarms FI/FSH-5*52 and -5*72 are located in the process vent lines connected to the 2 psig SCHe purge lines C and D. They monitor the flow from the 2 psig SCHe purge going to the process vent. The switch/alarm is non-safety class GS

  5. Low Power Wireless Smoke Alarm System in Home Fires

    Directory of Open Access Journals (Sweden)

    Juan Aponte Luis

    2015-08-01

    Full Text Available A novel sensing device for fire detection in domestic environments is presented. The fire detector uses a combination of several sensors that not only detect smoke, but discriminate between different types of smoke. This feature avoids false alarms and warns of different situations. Power consumption is optimized both in terms of hardware and software, providing a high degree of autonomy of almost five years. Data gathered from the device are transmitted through a wireless communication to a base station. The low cost and compact design provides wide application prospects.

  6. Statistical Study of False Alarms of Geomagnetic Storms

    DEFF Research Database (Denmark)

    Leer, Kristoffer; Vennerstrøm, Susanne; Veronig, A.

    . A subset of these halo CMEs did not cause a geomagnetic storm the following four days and have therefore been considered as false alarms. The properties of these events are investigated and discussed here. Their statistics are compared to the geo-effective CMEs. The ability to identify potential false......Coronal Mass Ejections (CMEs) are known to cause geomagnetic storms on Earth. However, not all CMEs will trigger geomagnetic storms, even if they are heading towards the Earth. In this study, front side halo CMEs with speed larger than 500 km/s have been identified from the SOHO LASCO catalogue...

  7. Olefins and chemical regulation in Europe: REACH.

    Science.gov (United States)

    Penman, Mike; Banton, Marcy; Erler, Steffen; Moore, Nigel; Semmler, Klaus

    2015-11-05

    REACH (Registration, Evaluation, Authorisation and Restriction of Chemicals) is the European Union's chemical regulation for the management of risk to human health and the environment (European Chemicals Agency, 2006). This regulation entered into force in June 2007 and required manufacturers and importers to register substances produced in annual quantities of 1000 tonnes or more by December 2010, with further deadlines for lower tonnages in 2013 and 2018. Depending on the type of registration, required information included the substance's identification, the hazards of the substance, the potential exposure arising from the manufacture or import, the identified uses of the substance, and the operational conditions and risk management measures applied or recommended to downstream users. Among the content developed to support this information were Derived No-Effect Levels or Derived Minimal Effect Levels (DNELs/DMELs) for human health hazard assessment, Predicted No Effect Concentrations (PNECs) for environmental hazard assessment, and exposure scenarios for exposure and risk assessment. Once registered, substances may undergo evaluation by the European Chemicals Agency (ECHA) or Member State authorities and be subject to requests for additional information or testing as well as additional risk reduction measures. To manage the REACH registration and related activities for the European olefins and aromatics industry, the Lower Olefins and Aromatics REACH Consortium was formed in 2008 with administrative and technical support provided by Penman Consulting. A total of 135 substances are managed by this group including 26 individual chemical registrations (e.g. benzene, 1,3-butadiene) and 13 categories consisting of 5-26 substances. This presentation will describe the content of selected registrations prepared for 2010 in addition to the significant post-2010 activities. Beyond REACH, content of the registrations may also be relevant to other European activities, for

  8. Caution and Warning Alarm Design and Evaluation for NASA CEV Auditory Displays: SHFE Information Presentation Directed Research Project (DRPP) report 12.07

    Science.gov (United States)

    Begault, Durand R.; Godfroy, Martine; Sandor, Aniko; Holden, Kritina

    2008-01-01

    The design of caution-warning signals for NASA s Crew Exploration Vehicle (CEV) and other future spacecraft will be based on both best practices based on current research and evaluation of current alarms. A design approach is presented based upon cross-disciplinary examination of psychoacoustic research, human factors experience, aerospace practices, and acoustical engineering requirements. A listening test with thirteen participants was performed involving ranking and grading of current and newly developed caution-warning stimuli under three conditions: (1) alarm levels adjusted for compliance with ISO 7731, "Danger signals for work places - Auditory Danger Signals", (2) alarm levels adjusted to an overall 15 dBA s/n ratio and (3) simulated codec low-pass filtering. Questionnaire data yielded useful insights regarding cognitive associations with the sounds.

  9. Associations between the self-reported frequency of hearing chemical alarms in theater and regional brain volume in Gulf War Veterans.

    Science.gov (United States)

    Chao, Linda L; Reeb, Rosemary; Esparza, Iva L; Abadjian, Linda R

    2016-03-01

    We previously reported evidence of reduced cortical gray matter (GM), white matter (WM), and hippocampal volume in Gulf War (GW) veterans with predicted exposure to low-levels of nerve agent according to the 2000 Khamisiyah plume model analysis. Because there is suggestive evidence that other nerve agent exposures may have occurred during the Gulf War, we examined the association between the self-reported frequency of hearing chemical alarms sound during deployment in the Gulf War and regional brain volume in GW veterans. Ninety consecutive GW veterans (15 female, mean age: 52±8years) participating in a VA-funded study underwent structural magnetic resonance imaging (MRI) on a 3T scanner. Freesurfer (version 5.1) was used to obtain regional measures of cortical GM, WM, hippocampal, and insula volume. Multiple linear regression was used to determine the association between the self-reported frequencies of hearing chemical alarms during the Gulf War and regional brain volume. There was an inverse association between the self-reported frequency of hearing chemical alarms sound and total cortical GM (adjusted p=0.007), even after accounting for potentially confounding demographic and clinical variables, the veterans' current health status, and other concurrent deployment-related exposures that were correlated with hearing chemical alarms. Post-hoc analyses extended the inverse relationship between the frequency of hearing chemical alarms to GM volume in the frontal (adjusted p=0.02), parietal (adjusted p=0.01), and occipital (adjusted p=0.001) lobes. In contrast, regional brain volumes were not significantly associated with predicted exposure to the Khamisiyah plume or with Gulf War Illness status defined by the Kansas or Centers for Disease Control and Prevention criteria. Many veterans reported hearing chemical alarms sound during the Gulf War. The current findings suggest that exposure to substances that triggered those chemical alarms during the Gulf War likely

  10. Stream Habitat Reach Summary - North Coast [ds63

    Data.gov (United States)

    California Natural Resource Agency — The shapefile is based on habitat unit level data summarized at the stream reach level. The database represents salmonid stream habitat surveys from 645 streams of...

  11. REACH and nanomaterials: current status

    International Nuclear Information System (INIS)

    Alessandrelli, Maria; Di Prospero Fanghella, Paola; Polci, Maria Letizia; Castelli, Stefano; Pettirossi, Flavio

    2015-01-01

    New challenges for regulators are emerging about a specific assessment and appropriate management of the potential risks of nanomaterials. In the framework of European legislation on chemicals, Regulation (EC) No. 1907/2006 REACH aims to ensure the safety of human health and the environment through the collection of information on the physico-chemical characteristics of the substances and on their profile (eco) toxicological and the identification of appropriate risk management linked to 'exposure to these substances without impeding scientific progress and the competitiveness of industry. In order to cover the current shortage of information on the safety of nanomaterials and tackle the acknowledged legal vacuum, are being a rich activities, carried out both by regulators both by stake holders, and discussions on the proposals for adapting the European regulatory framework for chemicals . The European Commission is geared to strengthen the REACH Regulation by means of updates of its annexes. The importance of responding to the regulatory requirements has highlighted the need for cooperation between European organizations, scientists and industries to promote and ensure the safe use of nanomaterials. [it

  12. Development of an integrated knowledge-base and its management tool for computerized alarm processing system

    International Nuclear Information System (INIS)

    Heo, Gyun Young; Choi, Seong Soo; Kim, Han Gon; Chang, Soon Heung

    1997-01-01

    For a long time, a number of alarm processing techniques have been researched to reduce the number of actuated alarms for operators to deal with effectively during the abnormal as well as the normal conditions. However, the strategy that the only systems with a few clear technologies should be used as a part of an alarm annunciation system has been adopted considering the effectiveness and the reliability in actual alarm processing systems. Therefore, alarm processing systems have difficult knowledge-base maintenance problems and limited expansion or enhancement defects. To solve these shortcomings, the integrated knowledge-base which can express the general information related to all the alarm processing techniques is proposed and its management tool, Knowledge Input Tool for Alarm (KIT-A) which can handle the data of the knowledge-base efficiently is developed. Since the integrated knowledge-base with KIT-A can manipulate all the alarm information without the modification of alarm processing system itself, it is expected to considerably advance the overall capability of maintenance and enhancement of the alarm processing systems

  13. Isolation of a pyrazine alarm pheromone component from the fire ant, Solenopsis invicta.

    Science.gov (United States)

    Vander Meer, Robert K; Preston, Catherine A; Choi, Man-Yeon

    2010-02-01

    Alarm pheromones in social insects are an essential part of a complex of pheromone interactions that contribute to the maintenance of colony integrity and sociality. The alarm pheromones of ants were among the first examples of animal pheromones identified, primarily because of the large amount of chemical produced and the distinctive responses of ants to the pheromone. However, the alarm pheromone of the fire ant, Solenopsis invicta, eluded identification for over four decades. We identified 2-ethyl-3,6-dimethylpyrazine as an alarm pheromone component of S. invicta. Worker fire ants detect the pyrazine alarm pheromone at 30 pg/ml, which is comparable to alarm pheromone sensitivities reported for other ant species. The source of this alarm pheromone are the mandibular glands, which, in fire ants, are not well developed and contain only about 300 pg of the compound, much less than the microgram quantities of alarm pheromones reported for several other ant species. Female and male sexuals and workers produce the pyrazine, which suggests that it may be involved in fire ant mating flight initiation, as well as the typical worker alarm response. This is the first report of 2-ethyl-3,6-dimethylpyrazine from a Solenopsis species and the first example of this alkaloid functioning as an alarm pheromone.

  14. Flights of fear: a mechanical wing whistle sounds the alarm in a flocking bird.

    Science.gov (United States)

    Hingee, Mae; Magrath, Robert D

    2009-12-07

    Animals often form groups to increase collective vigilance and allow early detection of predators, but this benefit of sociality relies on rapid transfer of information. Among birds, alarm calls are not present in all species, while other proposed mechanisms of information transfer are inefficient. We tested whether wing sounds can encode reliable information on danger. Individuals taking off in alarm fly more quickly or ascend more steeply, so may produce different sounds in alarmed than in routine flight, which then act as reliable cues of alarm, or honest 'index' signals in which a signal's meaning is associated with its method of production. We show that crested pigeons, Ocyphaps lophotes, which have modified flight feathers, produce distinct wing 'whistles' in alarmed flight, and that individuals take off in alarm only after playback of alarmed whistles. Furthermore, amplitude-manipulated playbacks showed that response depends on whistle structure, such as tempo, not simply amplitude. We believe this is the first demonstration that flight noise can send information about alarm, and suggest that take-off noise could provide a cue of alarm in many flocking species, with feather modification evolving specifically to signal alarm in some. Similar reliable cues or index signals could occur in other animals.

  15. ALMA Telescope Reaches New Heights

    Science.gov (United States)

    2009-09-01

    The ALMA (Atacama Large Millimeter/submillimeter Array) astronomical observatory took another step forward and upward, as one of its state-of-the-art antennas was carried for the first time to Chile's 16,500-foot-high plateau of Chajnantor on the back of a giant, custom-built transporter. The 40-foot-diameter antenna, weighing about 100 tons, was moved to ALMA's high-altitude Array Operations Site, where the extremely dry and rarefied air is ideal for observing the Universe. The conditions at the Array Operations Site on Chajnantor, while excellent for astronomy, are also very harsh. Only about half as much oxygen is available as at sea level, making it very difficult to work there. This is why ALMA's antennas are assembled and tested at the lower 9,500-foot altitude of the ALMA Operations Support Facility (OSF). It was from this relatively hospitable base camp that the ALMA antenna began its journey to the high Chajnantor site. "The successful transport of the first ALMA Antenna to the high site marks the start of the next phase of the project. Now that we are starting to move the ALMA antennas to the high site, the real work begins and the exciting part is just beginning," said Adrian Russell, North American ALMA Project Manager. The antenna's trip began when one of the two ALMA transporters lifted the antenna onto its back, carrying its heavy load along the 17-mile road from the Operations Support Facility up to the Array Operations Site. While the transporter is capable of speeds of up to 8 miles per hour when carrying an antenna, this first journey was made more slowly to ensure that everything worked as expected, taking about seven hours. The ALMA antennas use state-of-the-art technology, and are the most advanced submillimeter-wavelength antennas ever made. They are designed to operate fully exposed in the harsh conditions of the Array Operations Site, to survive strong winds and extreme temperatures, to point precisely enough that they could pick out a golf

  16. Honey Bee Inhibitory Signaling Is Tuned to Threat Severity and Can Act as a Colony Alarm Signal.

    Directory of Open Access Journals (Sweden)

    Ken Tan

    2016-03-01

    Full Text Available Alarm communication is a key adaptation that helps social groups resist predation and rally defenses. In Asia, the world's largest hornet, Vespa mandarinia, and the smaller hornet, Vespa velutina, prey upon foragers and nests of the Asian honey bee, Apis cerana. We attacked foragers and colony nest entrances with these predators and provide the first evidence, in social insects, of an alarm signal that encodes graded danger and attack context. We show that, like Apis mellifera, A. cerana possesses a vibrational "stop signal," which can be triggered by predator attacks upon foragers and inhibits waggle dancing. Large hornet attacks were more dangerous and resulted in higher bee mortality. Per attack at the colony level, large hornets elicited more stop signals than small hornets. Unexpectedly, stop signals elicited by large hornets (SS large hornet had a significantly higher vibrational fundamental frequency than those elicited by small hornets (SS small hornet and were more effective at inhibiting waggle dancing. Stop signals resulting from attacks upon the nest entrance (SS nest were produced by foragers and guards and were significantly longer in pulse duration than stop signals elicited by attacks upon foragers (SS forager. Unlike SS forager, SS nest were targeted at dancing and non-dancing foragers and had the common effect, tuned to hornet threat level, of inhibiting bee departures from the safe interior of the nest. Meanwhile, nest defenders were triggered by the bee alarm pheromone and live hornet presence to heat-ball the hornet. In A. cerana, sophisticated recruitment communication that encodes food location, the waggle dance, is therefore matched with an inhibitory/alarm signal that encodes information about the context of danger and its threat level.

  17. Honey Bee Inhibitory Signaling Is Tuned to Threat Severity and Can Act as a Colony Alarm Signal.

    Science.gov (United States)

    Tan, Ken; Dong, Shihao; Li, Xinyu; Liu, Xiwen; Wang, Chao; Li, Jianjun; Nieh, James C

    2016-03-01

    Alarm communication is a key adaptation that helps social groups resist predation and rally defenses. In Asia, the world's largest hornet, Vespa mandarinia, and the smaller hornet, Vespa velutina, prey upon foragers and nests of the Asian honey bee, Apis cerana. We attacked foragers and colony nest entrances with these predators and provide the first evidence, in social insects, of an alarm signal that encodes graded danger and attack context. We show that, like Apis mellifera, A. cerana possesses a vibrational "stop signal," which can be triggered by predator attacks upon foragers and inhibits waggle dancing. Large hornet attacks were more dangerous and resulted in higher bee mortality. Per attack at the colony level, large hornets elicited more stop signals than small hornets. Unexpectedly, stop signals elicited by large hornets (SS large hornet) had a significantly higher vibrational fundamental frequency than those elicited by small hornets (SS small hornet) and were more effective at inhibiting waggle dancing. Stop signals resulting from attacks upon the nest entrance (SS nest) were produced by foragers and guards and were significantly longer in pulse duration than stop signals elicited by attacks upon foragers (SS forager). Unlike SS forager, SS nest were targeted at dancing and non-dancing foragers and had the common effect, tuned to hornet threat level, of inhibiting bee departures from the safe interior of the nest. Meanwhile, nest defenders were triggered by the bee alarm pheromone and live hornet presence to heat-ball the hornet. In A. cerana, sophisticated recruitment communication that encodes food location, the waggle dance, is therefore matched with an inhibitory/alarm signal that encodes information about the context of danger and its threat level.

  18. Orthos, an alarm system for the ALICE DAQ operations

    Science.gov (United States)

    Chapeland, Sylvain; Carena, Franco; Carena, Wisla; Chibante Barroso, Vasco; Costa, Filippo; Denes, Ervin; Divia, Roberto; Fuchs, Ulrich; Grigore, Alexandru; Simonetti, Giuseppe; Soos, Csaba; Telesca, Adriana; Vande Vyvre, Pierre; von Haller, Barthelemy

    2012-12-01

    ALICE (A Large Ion Collider Experiment) is the heavy-ion detector studying the physics of strongly interacting matter and the quark-gluon plasma at the CERN LHC (Large Hadron Collider). The DAQ (Data Acquisition System) facilities handle the data flow from the detectors electronics up to the mass storage. The DAQ system is based on a large farm of commodity hardware consisting of more than 600 devices (Linux PCs, storage, network switches), and controls hundreds of distributed hardware and software components interacting together. This paper presents Orthos, the alarm system used to detect, log, report, and follow-up abnormal situations on the DAQ machines at the experimental area. The main objective of this package is to integrate alarm detection and notification mechanisms with a full-featured issues tracker, in order to prioritize, assign, and fix system failures optimally. This tool relies on a database repository with a logic engine, SQL interfaces to inject or query metrics, and dynamic web pages for user interaction. We describe the system architecture, the technologies used for the implementation, and the integration with existing monitoring tools.

  19. Orthos, an alarm system for the ALICE DAQ operations

    International Nuclear Information System (INIS)

    Chapeland, Sylvain; Carena, Franco; Carena, Wisla; Chibante Barroso, Vasco; Costa, Filippo; Divia, Roberto; Fuchs, Ulrich; Grigore, Alexandru; Simonetti, Giuseppe; Soos, Csaba; Telesca, Adriana; Vande Vyvre, Pierre; Von Haller, Barthelemy; Denes, Ervin

    2012-01-01

    ALICE (A Large Ion Collider Experiment) is the heavy-ion detector studying the physics of strongly interacting matter and the quark-gluon plasma at the CERN LHC (Large Hadron Collider). The DAQ (Data Acquisition System) facilities handle the data flow from the detectors electronics up to the mass storage. The DAQ system is based on a large farm of commodity hardware consisting of more than 600 devices (Linux PCs, storage, network switches), and controls hundreds of distributed hardware and software components interacting together. This paper presents Orthos, the alarm system used to detect, log, report, and follow-up abnormal situations on the DAQ machines at the experimental area. The main objective of this package is to integrate alarm detection and notification mechanisms with a full-featured issues tracker, in order to prioritize, assign, and fix system failures optimally. This tool relies on a database repository with a logic engine, SQL interfaces to inject or query metrics, and dynamic web pages for user interaction. We describe the system architecture, the technologies used for the implementation, and the integration with existing monitoring tools.

  20. Changes are detected - cameras and video systems are monitoring the plant site, only rarely giving false alarm

    International Nuclear Information System (INIS)

    Zeissler, H.

    1988-01-01

    The main purpose of automatic data acquisition and processing for monitoring goals is to relieve the security personnel from monotonous observation tasks. The novel video systems can be programmed to detect moving target alarm signals, or accept alarm-suppressing image changes. This allows an intelligent alarm evaluation for physical protection in industry, differentiating between real and false alarm signals. (orig.) [de

  1. Improving exposure scenario definitions within REACH

    DEFF Research Database (Denmark)

    Lee, Jihyun; Pizzol, Massimo; Thomsen, Marianne

    In recent years, the paradigm of chemical management system has changed from being toxicity oriented and media based to being risk oriented and receptor based. This trend is evident not only regarding environmental quality standards, but also for industrial chemical regulations. Political...... instruments to support a precautionary chemicals management system and to protect receptor’s health have also been increasing. Since 2007, the European Union adopted REACH (the Regulation on Registration, Evaluation, Authorisation and Restriction of Chemicals): REACH makes industry responsible for assessing...... and managing the risks posed by industrial chemicals and providing appropriate safety information to their users (EC, 2007). However, to ensure a high level of protection of human health and the environment, there is a need to consider ‘aggregate exposure’ including background exposures from environment which...

  2. Integrated built-in-test false and missed alarms reduction based on forward infinite impulse response & recurrent finite impulse response dynamic neural networks

    Science.gov (United States)

    Cui, Yiqian; Shi, Junyou; Wang, Zili

    2017-11-01

    Built-in tests (BITs) are widely used in mechanical systems to perform state identification, whereas the BIT false and missed alarms cause trouble to the operators or beneficiaries to make correct judgments. Artificial neural networks (ANN) are previously used for false and missed alarms identification, which has the features such as self-organizing and self-study. However, these ANN models generally do not incorporate the temporal effect of the bottom-level threshold comparison outputs and the historical temporal features are not fully considered. To improve the situation, this paper proposes a new integrated BIT design methodology by incorporating a novel type of dynamic neural networks (DNN) model. The new DNN model is termed as Forward IIR & Recurrent FIR DNN (FIRF-DNN), where its component neurons, network structures, and input/output relationships are discussed. The condition monitoring false and missed alarms reduction implementation scheme based on FIRF-DNN model is also illustrated, which is composed of three stages including model training, false and missed alarms detection, and false and missed alarms suppression. Finally, the proposed methodology is demonstrated in the application study and the experimental results are analyzed.

  3. A Knowledge-Based Approach to Automatic Detection of Equipment Alarm Sounds in a Neonatal Intensive Care Unit Environment.

    Science.gov (United States)

    Raboshchuk, Ganna; Nadeu, Climent; Jancovic, Peter; Lilja, Alex Peiro; Kokuer, Munevver; Munoz Mahamud, Blanca; Riverola De Veciana, Ana

    2018-01-01

    A large number of alarm sounds triggered by biomedical equipment occur frequently in the noisy environment of a neonatal intensive care unit (NICU) and play a key role in providing healthcare. In this paper, our work on the development of an automatic system for detection of acoustic alarms in that difficult environment is presented. Such automatic detection system is needed for the investigation of how a preterm infant reacts to auditory stimuli of the NICU environment and for an improved real-time patient monitoring. The approach presented in this paper consists of using the available knowledge about each alarm class in the design of the detection system. The information about the frequency structure is used in the feature extraction stage, and the time structure knowledge is incorporated at the post-processing stage. Several alternative methods are compared for feature extraction, modeling, and post-processing. The detection performance is evaluated with real data recorded in the NICU of the hospital, and by using both frame-level and period-level metrics. The experimental results show that the inclusion of both spectral and temporal information allows to improve the baseline detection performance by more than 60%.

  4. Summary of work carried out in the field of N4 plant series alarm processing

    International Nuclear Information System (INIS)

    Joly, J.-J.; Maurin, S.

    1990-01-01

    This document summarizes work executed or under way regarding the processing of new N4 plant series alarms. A review of the specific objectives is provided as well as a general description of the processing of these alarms. A description of the systems and prototypes developed is then provided: -autonomous N4 alarm supplementary processing system, -logic check expert system, -functional design check expert system based on qualitative system modeling

  5. Volatile compound diversity and conserved alarm behaviour in Triatoma dimidiata.

    Science.gov (United States)

    May-Concha, Irving; Rojas, Julio C; Cruz-López, Leopoldo; Ibarra-Cerdeña, Carlos N; Ramsey, Janine M

    2015-02-06

    Triatoma dimidiata (Latreille) is a key vector complex of Trypanosoma cruzi, etiologic agent of Chagas disease, as it spans North, Central, and South America. Although morphological and genetic studies clearly indicate existence of at least five clades within the species, there has been no robust or systematic revision, or appropriate nomenclature change for species within the complex. Three of the clades (haplogroups) are distributed in Mexico, and recent evidence attests to dispersal of clades across previously "presumed" monotypic geographic regions. Evidence of niche conservatism among sister species of this complex suggests that geographic dispersal is possible for non-sympatric populations, although no information is available on the behavioural aspects of potential interclade interactions, for instance whether differentiation of chemical signaling or response to these signals could impede communication among the haplogroups. Volatiles emitted by disturbed bugs, Brindley's (BGs), and metasternal (MGs) glands were identified using solid-phase micro-extraction (SPME) and gas chromatography coupled mass spectrometry (GC-MS). Volatile compounds emitted by BGs and MGs, and those secreted by disturbed nymphs and adults, of the three Mexican T. dimidiata haplogroups were tested for avoidance behaviour by conspecific nymphs and adults using an olfactometer. Triatoma dimidiata haplogroups all have three age-related alarm responses: absence of response by early stage nymphs, stage-specific response by 4-5th stage nymphs, and a shared 4-5th nymph and adult response to adult compounds. Disturbed bugs released 15 to 24 compounds depending on the haplogroup, among which were three pyrazines, the first report of these organoleptics in Triatominae. Isobutyric acid from BGs was the most abundant molecule in the response in all haplogroups, in addition to 15 (h1) to 21 (h2 and h3) MG compounds. Avoidance behaviour of disturbed bugs and volatiles emitted by BGs were haplogroup

  6. ALARM, Thermohydraulics of BWR with Jet Pumps During LOCA

    International Nuclear Information System (INIS)

    Araya, F.; Akimoto, M.

    1985-01-01

    1 - Nature of physical problem solved: ALARM-B2 which is an improved version of ALARM-B1 is a computer program to analyze thermo-hydraulic phenomena of BWR during a blowdown period under a large-break loss-of-coolant accident condition with special emphasis on the heat transfer phenomena in the core region. 2 - Method of solution: A so called volume-junction method is used to present fluid conservations. The primary system is divided into a number of special elements called 'control-volumes'. The system of partial differential equations describing fluid conservations for a stream-tube are integrated over a number of control volumes. The resulting set of simultaneous differential equations that is based on the assumptions of one-dimensional, homogeneous and thermal- equilibrium flow is linearized and solved for a small time increment by a simple explicit numerical technique. The one-dimensional heat conduction equations describing temperature profiles within solid material are written in finite difference forms which are linearized and solved by the Crank-Nicholson implicit method. In order to simulate the blowdown heat transfer phenomena, the code has correlation packages for heat transfer coefficient and critical heat flux. The heat generation in the core is given by a point reactor kinetics model with six groups of delayed neutrons and decay of eleven groups of fission products and actinides. The solution technique of the reactor kinetics is based on the Runge-Kutta method. ALARM-B2 has the models to simulate various components incorporated in BWRs such as jet pumps, recirculation pumps, steam separators, valves, and so on. The discharge and injection systems are modeled by leak and fill systems, respectively. 3 - Restrictions on the complexity of the problem: As this has been developed to simulate a blowdown thermo-hydraulic transient during a large break LOCA, users must pay attention when applying the code to any medium or small break LOCAs or to later phases

  7. Prepare 2 Learn: A mathematics intervention program for students at risk in Years 3 to 6 designed to help them reach expected level and become confident, responsible, independent mathematics learners

    OpenAIRE

    BERNADETTE MARY LONG

    2017-01-01

    This study reports on an intervention, Prepare 2 Learn, designed taking into account research literature and components of other successful mathematics programs. The research targeted students approximately 6 months behind the expected mathematics level for their year. The intervention consisted of four key components: building prerequisite knowledge of mathematical language, concepts, and skills to prepare students for their classroom mathematics; increasing fluency with mental computation; ...

  8. Advanced alarm system design and human performance: Guidance development and current research

    Energy Technology Data Exchange (ETDEWEB)

    O` Hara, J M [Brookhaven National Lab., Upton, NY (United States)

    1997-09-01

    This paper describes a research program sponsored by the U.S. Nuclear Regulatory Commission to address the human factors engineering (HFE) aspects of nuclear power plant alarm systems. The overall objective of the program is to develop HFE review guidance for advanced alarm systems. Guidance has been developed based on a broad base of technical and research literature. As part of the development effort, aspects of alarm system design for which the technical basis was insufficient to support guidance development were identified and prioritized. Research is currently underway to address the highest priority topics: alarm processing and display characteristics. (author). 29 refs, 2 figs.

  9. Accident alarm equipment for steam generator, especially liquid sodium heated steam generator

    International Nuclear Information System (INIS)

    Matal, O.; Jung, J.; Banovec, J.

    1982-01-01

    The alarm equipment consists of a system of sensors mounted onto the steam generator and its accessories. Each of the sensors is used for a different accident characteristic, such as the flow of sodium, the acoustic spectrum, the concentration of hydrogen in sodium. The system of sensors is connected to the common accident alarm system. The equipment will not issue the alarm signal if it receives a message from only one sensor, only when the message is confirmed from other sensors. This excludes false alarm. (M.D.)

  10. Advanced alarm system design and human performance: Guidance development and current research

    International Nuclear Information System (INIS)

    O'Hara, J.M.

    1997-01-01

    This paper describes a research program sponsored by the U.S. Nuclear Regulatory Commission to address the human factors engineering (HFE) aspects of nuclear power plant alarm systems. The overall objective of the program is to develop HFE review guidance for advanced alarm systems. Guidance has been developed based on a broad base of technical and research literature. As part of the development effort, aspects of alarm system design for which the technical basis was insufficient to support guidance development were identified and prioritized. Research is currently underway to address the highest priority topics: alarm processing and display characteristics. (author). 29 refs, 2 figs

  11. Development of Alarm System link Drawing for Operation Support for APR1400 Digital Main Control Room

    International Nuclear Information System (INIS)

    Kim, Ki-Hwan

    2016-01-01

    Digitalized MMI(Man-Machine Interface) including Digital Main Control Room(MCR) and digital I and C system was being applied for SKN 3 and 4 Nuclear Power Plant(NPP) and subsequent APR1400 NPP type. But, operators can not easily find instrument for alarm immediately. Therefore, Alarm system is required to easily find instrument for Alarm. For this implementation, we will plan system design considering design feature without affecting network load and CPU load. We have developed Alarm system link drawing for digital MCR. Operators of the digitalized MCR navigates from their consoles to the drawings related to the plant alarms and their instruments or the operation status. Such method gives cognitive load to the operators having to travel to different locations in finding the related information. Screen Sharing System, which is the fundamental technique for Drawing Interconnection Alarm System is close to completion, and it should be functionally tested and verified by the human factor engineering. For the actual application to the operating plants, the drawings to be interconnected to the alarms and the opinions from the operators/maintenance departments for designating alarm number should be surveyed, Also, another function that allows the access to the alarm related drawings not only from the MCR but also from the other offices

  12. Development of Alarm System link Drawing for Operation Support for APR1400 Digital Main Control Room

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Ki-Hwan [KHNP CRI, Daejeon (Korea, Republic of)

    2016-10-15

    Digitalized MMI(Man-Machine Interface) including Digital Main Control Room(MCR) and digital I and C system was being applied for SKN 3 and 4 Nuclear Power Plant(NPP) and subsequent APR1400 NPP type. But, operators can not easily find instrument for alarm immediately. Therefore, Alarm system is required to easily find instrument for Alarm. For this implementation, we will plan system design considering design feature without affecting network load and CPU load. We have developed Alarm system link drawing for digital MCR. Operators of the digitalized MCR navigates from their consoles to the drawings related to the plant alarms and their instruments or the operation status. Such method gives cognitive load to the operators having to travel to different locations in finding the related information. Screen Sharing System, which is the fundamental technique for Drawing Interconnection Alarm System is close to completion, and it should be functionally tested and verified by the human factor engineering. For the actual application to the operating plants, the drawings to be interconnected to the alarms and the opinions from the operators/maintenance departments for designating alarm number should be surveyed, Also, another function that allows the access to the alarm related drawings not only from the MCR but also from the other offices.

  13. Multi-parameter vital sign database to assist in alarm optimization for general care units.

    Science.gov (United States)

    Welch, James; Kanter, Benjamin; Skora, Brooke; McCombie, Scott; Henry, Isaac; McCombie, Devin; Kennedy, Rosemary; Soller, Babs

    2016-12-01

    Continual vital sign assessment on the general care, medical-surgical floor is expected to provide early indication of patient deterioration and increase the effectiveness of rapid response teams. However, there is concern that continual, multi-parameter vital sign monitoring will produce alarm fatigue. The objective of this study was the development of a methodology to help care teams optimize alarm settings. An on-body wireless monitoring system was used to continually assess heart rate, respiratory rate, SpO 2 and noninvasive blood pressure in the general ward of ten hospitals between April 1, 2014 and January 19, 2015. These data, 94,575 h for 3430 patients are contained in a large database, accessible with cloud computing tools. Simulation scenarios assessed the total alarm rate as a function of threshold and annunciation delay (s). The total alarm rate of ten alarms/patient/day predicted from the cloud-hosted database was the same as the total alarm rate for a 10 day evaluation (1550 h for 36 patients) in an independent hospital. Plots of vital sign distributions in the cloud-hosted database were similar to other large databases published by different authors. The cloud-hosted database can be used to run simulations for various alarm thresholds and annunciation delays to predict the total alarm burden experienced by nursing staff. This methodology might, in the future, be used to help reduce alarm fatigue without sacrificing the ability to continually monitor all vital signs.

  14. INEL test reactor facility alarms: descriptions, technical specifications, and modification procedure

    International Nuclear Information System (INIS)

    Potash, L.M.; Boone, M.P.

    1980-04-01

    This report identifies standards, procedures, and practices which will affect any attempt to integrate or introduce human engineering principles into nuclear power plant alarm systems. Additional information concerning type of signal used, expected reaction, type of sensor, etc., is presented because of its relevance to future work on alarm system integration. The INEL test reactors were studied. Interviews were conducted with operators, designers, and management personnel. Additional information was obtained from available documentation. Only fire-alarm systems, and to a lesser extent, criticality alarms, have detailed industry-wide standards. One general standard has been written for control-room annunciators

  15. Non-specific alarm calls trigger mobbing behavior in Hainan gibbons (Nomascus hainanus).

    Science.gov (United States)

    Deng, Huaiqing; Gao, Kai; Zhou, Jiang

    2016-09-30

    Alarm calls are important defensive behaviors. Here, we report the acoustic spectrum characteristics of alarm calls produced by Hainan gibbons (Nomascus hainanus) inhabiting Bawangling National Nature Reserve in Hainan, China. Analysis of call data collected from 2002-2014 shows that alarm calls are emitted by all family group members, except infants. Alarm behavior included simple short alarming calls (7-10 min) followed by longer variable-frequency mobbing calls lasting 5-12 min. The duration of individual alarming and mobbing calls was 0.078 ± 0.014 s and 0.154 ± 0.041 s at frequency ranges of 520-1000 Hz and 690-3920 Hz, respectively. Alarming call duration was positively associated with group size. The alarm calls can trigger mobbing behavior in Hainan gibbons; this is a defense way of social animals, and first report among the primates' species. The system of vocal alarm behavior described in this critically endangered species is simple and effective.

  16. Resolving MC and A alarms from process monitoring in a fuel fabrication facility

    International Nuclear Information System (INIS)

    Smith, B.W.; Razvi, J.

    1984-01-01

    Process monitoring data can be used for generating material loss estimates. The intent of using process control data is to enhance nuclear material control and accounting for the timely detection and resolution of discrepancies. The purpose of an alarm resolution system is to distinguish between system errors and an actual loss of nuclear material. A study has been performed to develop and test a site-specific set of alarm resolution procedures. The results of the study are described and include the frequency of alarms, the causes of alarms, the type of resolution, and the modeling of loss estimates

  17. A Human Factors Perspective on Alarm System Research and Development 2000 to 2010

    Energy Technology Data Exchange (ETDEWEB)

    Curt Braun; John Grimes; Eric Shaver; Ronald Boring (Principal Investigator)

    2011-09-01

    By definition, alarms serve to notify human operators of out-of-parameter conditions that could threaten equipment, the environment, product quality and, of course, human life. Given the complexities of industrial systems, human machine interfaces, and the human operator, the understanding of how alarms and humans can best work together to prevent disaster is continually developing. This review examines advances in alarm research and development from 2000 to 2010 and includes the writings of trade professionals, engineering and human factors researchers, and standards organizations with the goal of documenting advances in alarms system design, research, and implementation.

  18. Reaching for the red planet

    Science.gov (United States)

    David, L

    1996-05-01

    The distant shores of Mars were reached by numerous U.S. and Russian spacecraft throughout the 1960s to mid 1970s. Nearly 20 years have passed since those successful missions which orbited and landed on the Martian surface. Two Soviet probes headed for the planet in July, 1988, but later failed. In August 1993, the U.S. Mars Observer suddenly went silent just three days before it was to enter orbit around the planet and was never heard from again. In late 1996, there will be renewed activity on the launch pads with three probes departing for the red planet: 1) The U.S. Mars Global Surveyor will be launched in November on a Delta II rocket and will orbit the planet for global mapping purposes; 2) Russia's Mars '96 mission, scheduled to fly in November on a Proton launcher, consists of an orbiter, two small stations which will land on the Martian surface, and two penetrators that will plow into the terrain; and finally, 3) a U.S. Discovery-class spacecraft, the Mars Pathfinder, has a December launch date atop a Delta II booster. The mission features a lander and a microrover that will travel short distances over Martian territory. These missions usher in a new phase of Mars exploration, setting the stage for an unprecedented volley of spacecraft that will orbit around, land on, drive across, and perhaps fly at low altitudes over the planet.

  19. City Reach Code Technical Support Document

    Energy Technology Data Exchange (ETDEWEB)

    Athalye, Rahul A. [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Chen, Yan [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Zhang, Jian [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Liu, Bing [Pacific Northwest National Lab. (PNNL), Richland, WA (United States); Frankel, Mark [New Buildings Inst., Portland, OR (United States); Lyles, Mark [New Buildings Inst., Portland, OR (United States)

    2017-10-31

    This report describes and analyzes a set of energy efficiency measures that will save 20% energy over ASHRAE Standard 90.1-2013. The measures will be used to formulate a Reach Code for cities aiming to go beyond national model energy codes. A coalition of U.S. cities together with other stakeholders wanted to facilitate the development of voluntary guidelines and standards that can be implemented in stages at the city level to improve building energy efficiency. The coalition's efforts are being supported by the U.S. Department of Energy via Pacific Northwest National Laboratory (PNNL) and in collaboration with the New Buildings Institute.

  20. Development of semiconductor radiation sensors for portable alarm-dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Y. K.; Moon, B. S.; Chung, C. E.; Hong, S. B.; Kim, J. Y.; Kim, J. B.; Han, S. H.; Lee, W. G. [Korea Atomic Energy Research Institute, Taejeon (Korea)

    2001-01-01

    We studied Semiconductor Radiation Sensors for Portable Alarm-Dosimeter. We calculated response functions for gamma energy 0.021, 0.122, 0.662, 0.835, 1.2 MeV using EGS4 codes. When we measured at various distance from source to detector, the detection efficiency of Si semiconductor detector was better than that of GM tube. The linear absorption coefficients of steel and aluminum plate were measured. These experimental results of the response of detector for intensity of radiation field coincide to the theoretical expectation. The count value of Si detector was changed with changing thickness of steel as changing threshold voltage of discriminator, and the linear absorption coefficient increased with increasing threshold voltage. Radiation detection efficiency shows difference at each threshold voltage condition. This results coincided to the theoretical simulation. 33 refs., 27 figs., 8 tabs. (Author)

  1. A new fire alarm system for electrical installations

    CERN Document Server

    Pietersen, A H

    1978-01-01

    Fires in electrical installations are considered to develop in four phases - initiation, smouldering, flame formation and heat development. Cables are among the more sensitive components, with working temperatures around 50 degrees C and fire detection at 70 degrees C. Conventional alarms include smoke detectors. The new technique described uses microcapsules containing powder forming a gas of the Freon type after diffusion. A typical microcapsule loses 4% per year and has a natural life of 10 years. Fabrication methods are described. Detection is by gas concentration, with a sensitivity of 1 to 10 ppm, or by acoustic methods with microphones to pick up the sound of fractures. Pressure/temperature characteristics of various types of Freon mixtures commercially available are given in graphical form.

  2. Design of remote control alarm system by microwave detection

    Science.gov (United States)

    Wang, Junli

    2018-04-01

    A microwave detection remote control alarm system is designed, which is composed of a Microwave detectors, a radio receiving/transmitting module and a digital encoding/decoding IC. When some objects move into the surveillance area, microwave detectors will generate a control signal to start transmitting system. A radio control signal will be spread by the transmitting module, once the signal can be received, and it will be disposed by some circuits, arousing some voices that awake the watching people. The whole device is a modular configuration, it not only has some advantage of frequency stable, but also reliable and adjustment-free, and it is suitable for many kinds of demands within the distance of 100m.

  3. Definition of areas requiring criticality alarm annunciation and emergency control

    International Nuclear Information System (INIS)

    Hobson, J.M.

    1988-01-01

    The design of fissile material handling at British Nuclear Fuels plc requires the provision of a criticality incident detection system unless a specific case for omission can be formally made. Where such systems are provided, the 100 mSv contour resulting from a reference criticality incident must be restricted to an area of administrative control within which it is reasonably practicable to provide alarm annunciation and for which emergency arrangements can be defined. For typical reprocessing plant applications, the definition of these areas, and their restriction by provision of shielding where necessary, potentially requires a very large number of three dimensional neutron transport calculations in complex geometries. However, by considering the requirements and nature of this assessment, simple generic methods have been developed and justified. Consequently, rapid and inexpensive assessments of control areas can be carried out

  4. Soft real-time alarm messages for ATLAS TDAQ

    Science.gov (United States)

    Darlea, G.; Al Shabibi, A.; Martin, B.; Lehmann Miotto, G.

    2010-05-01

    The ATLAS TDAQ network consists of three separate Ethernet-based networks (Data, Control and Management) with over 2000 end-nodes. The TDAQ system has to be aware of the meaningful network failures and events in order for it to take effective recovery actions. The first stage of the process is implemented with Spectrum, a commercial network management tool. Spectrum detects and registers all network events, then it publishes the information via a CORBA programming interface. A gateway program (called NSG—Network Service Gateway) connects to Spectrum through CORBA and exposes to its clients a Java RMI interface. This interface implements a callback mechanism that allows the clients to subscribe for monitoring "interesting" parts of the network. The last stage of the TDAQ network monitoring tool is implemented in a module named DNC (DAQ to Network Connection), which filters the events that are to be reported to the TDAQ system: it subscribes to the gateway only for the machines that are currently active in the system and it forwards only the alarms that are considered important for the current TDAQ data taking session. The network information is then synthesized and presented in a human-readable format. These messages can be further processed either by the shifter who is in charge, the network expert or the Online Expert System. This article aims to describe the different mechanisms of the chain that transports the network events to the front-end user, as well as the constraints and rules that govern the filtering and the final format of the alarm messages.

  5. Soft real-time alarm messages for ATLAS TDAQ

    International Nuclear Information System (INIS)

    Darlea, G.; Al Shabibi, A.; Martin, B.; Lehmann Miotto, G.

    2010-01-01

    The ATLAS TDAQ network consists of three separate Ethernet-based networks (Data, Control and Management) with over 2000 end-nodes. The TDAQ system has to be aware of the meaningful network failures and events in order for it to take effective recovery actions. The first stage of the process is implemented with Spectrum, a commercial network management tool. Spectrum detects and registers all network events, then it publishes the information via a CORBA programming interface. A gateway program (called NSG-Network Service Gateway) connects to Spectrum through CORBA and exposes to its clients a Java RMI interface. This interface implements a callback mechanism that allows the clients to subscribe for monitoring 'interesting' parts of the network. The last stage of the TDAQ network monitoring tool is implemented in a module named DNC (DAQ to Network Connection), which filters the events that are to be reported to the TDAQ system: it subscribes to the gateway only for the machines that are currently active in the system and it forwards only the alarms that are considered important for the current TDAQ data taking session. The network information is then synthesized and presented in a human-readable format. These messages can be further processed either by the shifter who is in charge, the network expert or the Online Expert System. This article aims to describe the different mechanisms of the chain that transports the network events to the front-end user, as well as the constraints and rules that govern the filtering and the final format of the alarm messages.

  6. Soft real-time alarm messages for ATLAS TDAQ

    Energy Technology Data Exchange (ETDEWEB)

    Darlea, G., E-mail: georgiana.lavinia.darlea@cern.c [CERN, Geneva (Switzerland); Al Shabibi, A.; Martin, B.; Lehmann Miotto, G. [CERN, Geneva (Switzerland)

    2010-05-21

    The ATLAS TDAQ network consists of three separate Ethernet-based networks (Data, Control and Management) with over 2000 end-nodes. The TDAQ system has to be aware of the meaningful network failures and events in order for it to take effective recovery actions. The first stage of the process is implemented with Spectrum, a commercial network management tool. Spectrum detects and registers all network events, then it publishes the information via a CORBA programming interface. A gateway program (called NSG-Network Service Gateway) connects to Spectrum through CORBA and exposes to its clients a Java RMI interface. This interface implements a callback mechanism that allows the clients to subscribe for monitoring 'interesting' parts of the network. The last stage of the TDAQ network monitoring tool is implemented in a module named DNC (DAQ to Network Connection), which filters the events that are to be reported to the TDAQ system: it subscribes to the gateway only for the machines that are currently active in the system and it forwards only the alarms that are considered important for the current TDAQ data taking session. The network information is then synthesized and presented in a human-readable format. These messages can be further processed either by the shifter who is in charge, the network expert or the Online Expert System. This article aims to describe the different mechanisms of the chain that transports the network events to the front-end user, as well as the constraints and rules that govern the filtering and the final format of the alarm messages.

  7. A proactive alarm reduction method and its human factors validation test for a main control room for SMART

    International Nuclear Information System (INIS)

    Jang, Gwi-sook; Suh, Sang-moon; Kim, Sa-kil; Suh, Yong-suk; Park, Je-yun

    2013-01-01

    Highlights: ► A proactive alarm reduction method improves effectiveness on the alarm reduction. ► The method suppresses alarms based on the ECA rules and facts for the alarm reduction under an alarm flood situation. ► The alarm reduction logics are supplemented to a high hit ratio of the reduction logics during on-line operations. ► The method is validated by human factors validation test based on regulatory requirements. -- Abstract: Conventional alarm systems tend to overwhelm operators during a transient because of a large number of nearly simultaneous annunciator activations with varying degrees of relevance to operator tasks. Thus alarm processing techniques have developed to support operators in coping with the volume of alarms, to identify which alarms are significant, and to reduce the need for operators to infer the plant conditions. This paper proposes a proactive alarm reduction method for SMART (System-integrated Modular Advanced ReacTor) whereby based on the contents of the past operating effects alarm reduction is carried out during the next transient. We designed and implemented the proactive alarm reduction system and constructed the environment for the human factors validation test. Also, eight subjects actually working in a nuclear power plant (NPP) tested the practical effectiveness of the proposed proactive alarm reduction method according to the procedure of human factors validation test under a dynamic simulation of a partial scope for an NPP.

  8. Decreasing the false alarm rate of arrhythmias in intensive care using a machine learning approach

    NARCIS (Netherlands)

    Eerikäinen, L.M.; Vanschoren, J.; Rooijakkers, M.J.; Vullings, R.; Aarts, R.M.

    2015-01-01

    We present a novel algorithm for classifying true and false alarms of five life-threatening arrhythmias in intensive care. This algorithm was entered in the PhysioNet/Computing in Cardiology Challenge 2015 Reducing False Arrhythmia Alarms in the ICU. The algorithm performs a binary classification of

  9. Design for safety : A new service for alarming and informing the population in case of emergency

    NARCIS (Netherlands)

    Jagtman, H.M.

    2012-01-01

    In case of emergencies the population in danger should be alarmed so individuals can take action to get or remain out of danger. For alarming the population the means available are limited. Many countries have outdoor sirens. They operability however is limited since the siren has only one implicit

  10. Use of alarm features in referral of febrile children to the emergency department : an observational study

    NARCIS (Netherlands)

    van Ierland, Yvette; Elshout, Gijs; Moll, Henritte A.; Nijman, Ruud G.; Vergouwe, Yvonne; van der Lei, Johan; Berger, Marjolein Y.; Oostenbrink, Rianne

    Background The diagnostic value of alarm features of serious infections in low prevalence settings is unclear. Aim To explore to what extent alarm features play a role in referral to the emergency department (ED) by GPs who face a febrile child during out-of-hours care. Design and setting

  11. 40 CFR 264.34 - Access to communications or alarm system.

    Science.gov (United States)

    2010-07-01

    ... FACILITIES Preparedness and Prevention § 264.34 Access to communications or alarm system. (a) Whenever... 40 Protection of Environment 25 2010-07-01 2010-07-01 false Access to communications or alarm system. 264.34 Section 264.34 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) SOLID...

  12. 40 CFR 265.34 - Access to communications or alarm system.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 25 2010-07-01 2010-07-01 false Access to communications or alarm system. 265.34 Section 265.34 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) SOLID..., STORAGE, AND DISPOSAL FACILITIES Preparedness and Prevention § 265.34 Access to communications or alarm...

  13. Writing short alarm messages : A matter of education, training and practice

    NARCIS (Netherlands)

    Jagtman, H.M.

    2013-01-01

    The Netherlands has a new tool to alarm and inform the population in case of crises. NL-Alert can simultaneously draw people’s attention and explain the crisis matter. The use requires determining of both the impact area and the alarm text. This raises questions about required knowledge and

  14. The use of simulation in the development of human factors guidelines for alarm systems

    International Nuclear Information System (INIS)

    O'Hara, J.; Brown, W.S.; Wachtel, J.; Persensky, J.

    1997-01-01

    This paper describes a research program sponsored by the US Nuclear Regulatory Commission to address the human factors engineering (HFE) aspects of nuclear power plant alarm systems. The overall objective of the program is to develop HFE review guidance for advanced alarm systems. As part of this program, guidance was developed based on a broad review and analysis of technical and research literature. In the course of guidance development, aspects of alarm system design for which the technical basis was insufficient to support guidance developed were identified. Experimental research is currently underway to address the highest priority topics: alarm processing and display characteristics. This paper provides an overview of the approach to guidance development and discusses the role of simulation in the development approach. Finally, the current simulator-based experiment is described to illustrate how the alarm system design features are being studied

  15. Application of neural networks to multiple alarm processing and diagnosis in nuclear power plants

    International Nuclear Information System (INIS)

    Cheon, Se Woo; Chang Soon Heung; Chung, Hak Yeong

    1992-01-01

    This paper presents feasibility studies of multiple alarm processing and diagnosis using neural networks. The back-propagation neural network model is applied to the training of multiple alarm patterns for the identification of failure in a reactor coolant pump (RCP) system. The general mapping capability of the neural network enables to identify a fault easily. The case studies are performed with emphasis on the applicability of the neural network to pattern recognition problems. It is revealed that the neural network model can identify the cause of multiple alarms properly, even when untrained or sensor-failed alarm symptoms are given. It is also shown that multiple failures are easily identified using the symptoms of multiple alarms

  16. Stimulus-response time to alarms of the intra-aortic balloon pump: safe care practices

    Directory of Open Access Journals (Sweden)

    Andrezza Serpa Franco

    Full Text Available ABSTRACT Objective: To characterize the sound alarms of the Intra-Aortic Balloon Pump (IABP during aortic counterpulsation therapy; to measure the stimulus-response time of the team to these; and to discuss the implications of increasing this time for patient safety from the alarm fatigue perspective. Method: This is an observational and descriptive study with quantitative and qualitative approach, case study type, carried out in a Cardiac Surgical Intensive Care Unit. Results: The most audible IABP alarm was the one of high priority increased-reduced diastolic blood pressure. The stimulus-response time was 33.9 seconds on average. Conclusion: Managing the alarms of these equipment is essential to minimize the occurrence of the alarm fatigue phenomenon and to offer a safer assistance to patients who rely on this technology.

  17. Expert systems and microwave communication systems alarms processing: A feasibility study

    Energy Technology Data Exchange (ETDEWEB)

    Tonn, B.; Goeltz, R.; Purucker, S.

    1987-07-01

    This report presents the results of a feasibility study conducted by Oak Ridge National Laboratory (ORNL) for the Bonneville Power Administration concerning the applicability of Artificial Intelligence (AI) technology to process alarms associated with Bonneville's Microwave Communication System (MCS). Specifically, the discussion focuses on the characteristics of a prototype expert system/database management system (DBMS) configuration capable of intelligently processing alarms, efficiently storing alarm-based historical data, and providing analysis and reporting tools. Such a system has the potential to improve response to critical alarms, increase the information content of a large volume of complicated data, free operators from performing routine analysis, and provide alarm information to operators, field personnel, and management through queries and automatically produced reports.

  18. Adaptive mixed reality rehabilitation improves quality of reaching movements more than traditional reaching therapy following stroke.

    Science.gov (United States)

    Duff, Margaret; Chen, Yinpeng; Cheng, Long; Liu, Sheng-Min; Blake, Paul; Wolf, Steven L; Rikakis, Thanassis

    2013-05-01

    Adaptive mixed reality rehabilitation (AMRR) is a novel integration of motion capture technology and high-level media computing that provides precise kinematic measurements and engaging multimodal feedback for self-assessment during a therapeutic task. We describe the first proof-of-concept study to compare outcomes of AMRR and traditional upper-extremity physical therapy. Two groups of participants with chronic stroke received either a month of AMRR therapy (n = 11) or matched dosing of traditional repetitive task therapy (n = 10). Participants were right handed, between 35 and 85 years old, and could independently reach to and at least partially grasp an object in front of them. Upper-extremity clinical scale scores and kinematic performances were measured before and after treatment. Both groups showed increased function after therapy, demonstrated by statistically significant improvements in Wolf Motor Function Test and upper-extremity Fugl-Meyer Assessment (FMA) scores, with the traditional therapy group improving significantly more on the FMA. However, only participants who received AMRR therapy showed a consistent improvement in kinematic measurements, both for the trained task of reaching to grasp a cone and the untrained task of reaching to push a lighted button. AMRR may be useful in improving both functionality and the kinematics of reaching. Further study is needed to determine if AMRR therapy induces long-term changes in movement quality that foster better functional recovery.

  19. Replacement of the criticality accident alarm system in the Tokai reprocessing plant

    International Nuclear Information System (INIS)

    Sanada, Yukihisa; Momose, Takumaro; Suzuki, Kei; Kawai, Keiichi

    2008-01-01

    A Criticality Accident Alarm System (CAAS) was installed as part of criticality safety management for use in reducing the radiation workers could be exposed to in the rare case of a criticality accident. The initial CAAS version was installed the Tokai Reprocessing Plant (TRP) in the 1980s. It includes units that can detect gamma-rays or neutron-rays released in criticality accidents (CADs), one of which consists of three plastic scintillation gamma detectors and three solid state neutron detectors with fissile material, and in being highly reliable utilizes the 2 out of 3 voting system. The purpose of this study is to give the design principles and procedures for determining the adequate relocation of the CADs within the TRP. The optimal places for the CADs to be relocated to were determined using a conservative evaluation method. Firstly, equipment needing to be monitored for criticality accidents was selected with consideration given to the risk of excessive exposure to workers. Secondly, the detection threshold of a minimum accident was set to be an increase in power of 10 15 fissions/s occurring within a rise-time of between 0.5 ms and 1 s. The sum of neutron and gamma doses of a minimum accident (10 15 fissions) was 0.3 Gy at an unshielded distance of 1 m. Finally, doses at where the CADs were installed were evaluated using parameters calculated with MCNP and ANISN. As a result, the alarm trip level of both the gamma detector and the neutron detector being set at 2.0 mGy/h enabled minimum criticality accidents to be conservatively detected. These results were then applied to the new CAD positions. (author)

  20. Minimum detectable activity and false alarm rate relationships for alpha continuous air monitors

    International Nuclear Information System (INIS)

    Hoover, M.D.; Newton, G.J.

    1994-01-01

    The U.S. Department of Energy rule for Occupational Radiation Protection (10 CFR Part 835, December 1993) and the DOE Radiological Control Manual (the RCM) (DOE/EH-0256T, Rev. 1, April 1994) require the use of continuous air monitors (CAMs) in normally occupied areas where an individual is likely to be exposed to a concentration of airborne radioactivity exceeding the derived air concentration (DAC) or where there is a need to alert potentially exposed individuals to unexpected increases in airborne radioactivity levels. The DAC is the airborne concentration that equals the annual limit on intake divided by the volume of air breathed by an average worker for a working year of 2000 h (assuming a breathing volume of 2400 m 3 ). It is equivalent to the airborne concentration to which a worker could be exposed for an entire working year (2000 h) without exceeding the allowable annual limit on intake. The rule and the RCM further require that real-time air monitors have an alarm capability and sufficient sensitivity to alert potentially exposed individuals that immediate action is necessary in order to minimize or terminate inhalation exposures. The RCM also recommends that real-time air monitors should be capable of measuring 1 DAC when averaged over 8 h (8 DAC-h) under laboratory conditions. In response to these recommendations, we are developing procedures for determining the basic sensitivity of alpha CAMs under laboratory conditions and for documenting practical alarm set points for routine use of CAMs under a range of radon and thoron concentrations

  1. Differences in alarm events between disposable and reusable electrocardiography lead wires.

    Science.gov (United States)

    Albert, Nancy M; Murray, Terri; Bena, James F; Slifcak, Ellen; Roach, Joel D; Spence, Jackie; Burkle, Alicia

    2015-01-01

    Disposable electrocardiographic lead wires (ECG-LWs) may not be as durable as reusable ones. To examine differences in alarm events between disposable and reusable ECG-LWs. Two cardiac telemetry units were randomized to reusable ECG-LWs, and 2 units alternated between disposable and reusable ECG-LWs for 4 months. A remote monitoring team, blinded to ECG-LW type, assessed frequency and type of alarm events by using total counts and rates per 100 patient days. Event rates were compared by using generalized linear mixed-effect models for differences and noninferiority between wire types. In 1611 patients and 9385.5 patient days of ECG monitoring, patient characteristics were similar between groups. Rates of alarms for no telemetry, leads fail, or leads off were lower in disposable ECG-LWs (adjusted relative risk [95% CI], 0.71 [0.53-0.96]; noninferiority P < .001; superiority P = .03) and monitoring (artifact) alarms were significantly noninferior (adjusted relative risk [95% CI]: 0.88, [0.62-1.24], P = .02; superiority P = .44). No between-group differences existed in false or true crisis alarms. Disposable ECG-LWs were noninferior to reusable ECG-LWs for all false-alarm events (N [rate per 100 patient days], disposable 2029 [79.1] vs reusable 6673 [97.9]; adjusted relative risk [95% CI]: 0.81 [0.63-1.06], P = .002; superiority P = .12.) Disposable ECG-LWs with patented push-button design had superior performance in reducing alarms created by no telemetry, leads fail, or leads off and significant noninferiority in all false-alarm rates compared with reusable ECG-LWs. Fewer ECG alarms may save nurses time, decrease alarm fatigue, and improve patient safety. ©2015 American Association of Critical-Care Nurses.

  2. Behavioral responses of Pacific lamprey to alarm cues

    Science.gov (United States)

    Porter, Laurie L.; Hayes, Michael C.; Jackson, Aaron D.; Burke, Brian J.; Moser, Mary L.; Wagner, R. Steven

    2017-01-01

    Pacific lamprey (Entosphenus tridentatus), an anadromous ectoparasite, faces several challenges during adult migration to spawning grounds. Developing methods to address these challenges is critical to the success of ongoing conservation efforts. The challenges are diverse, and include anthropogenic alterations to the ecosystem resulting in loss of habitat, impassable barriers such as dams, climate change impacts, and altered predator fields. We conducted a behavioral study to understand how adult migrating Pacific lamprey respond to potential alarm cues: White Sturgeon (Acipenser transmontanus), human saliva, decayed Pacific lamprey, and river otter (Lontra canadensis). Research has shown that some species of lamprey can be guided to a location using odors and similar cues may be useful as a management tool for Pacific lamprey. Experiments were conducted over 2 nights and measured the number of entries (count) and duration of time spent (occupancy) by adult lamprey in each arm of a two-choice maze. During the first night, no odor was added to test for selection bias between arms. During the second night odor was added to one arm of the maze. Contrary to expectations, lamprey were significantly attracted to the river otter odor in both count and occupancy. No significant differences were found in the response of lamprey to the other three odors. Results from this study indicate that Pacific lamprey do respond to some odors; however, additional tests are necessary to better identify the types of odors and concentrations that elicit a repeatable response.

  3. Cancer prevention and control: alarming challenges in China.

    Science.gov (United States)

    Bode, Ann M; Dong, Zigang; Wang, Hongyang

    2016-03-01

    China is geographically the third largest country in the world and the most populated low-to-middle-income country. Cancer incidence and mortality rates for some cancers in the USA and European countries have steadily decreased over the last decades, whereas the incidence and mortality of certain cancers in China have been increasing at an alarming speed. Rapid industrialization and urbanization in China have been accompanied by incredible changes in lifestyle and environment combined with an aging population. Mortality caused by lung, colorectal and breast cancers has been steadily increasing, whereas cancer mortality from gastric, esophageal and cervical tumors has tended to decrease. Similar to what has occurred in the United States, unhealthy lifestyles in China, including heavy smoking and poor diet combined with pollution, have contributed to increased cancer risk. China is facing many challenges in cancer treatment and prevention for the general population. The major areas that need to be addressed in the control of cancer in China include cancers associated with environmental pollution, tobacco use, occupational carcinogens, infection, excessive alcohol consumption, dietary deficiencies and obesity. In this perspective, we review the problems in each area and suggest ideas for future directions in cancer research and strategies and actions to reduce the incidence of cancer in China.

  4. Operational Reach: Is Current Army Doctrine Adequate?

    National Research Council Canada - National Science Library

    Heintzelman, Scott

    2003-01-01

    The term operational reach, an element of operational design, is new to U.S. Army doctrine. Operational reach is not found in the previous edition of the Army's basic operational doctrine, Field Manual...

  5. English Learners: Reaching the Highest Level of English Literacy.

    Science.gov (United States)

    Garcia, Gilbert C., Ed.

    This collection of papers examines the critical literacy development of English learners, focusing on English reading instruction in an immersion setting, English language development, and cultural issues pertaining to English learners in and out of the classroom. The 16 papers include the following: (1) "Reading and the Bilingual Student: Fact…

  6. Alarm pheromone does not modulate 22-kHz calls in male rats.

    Science.gov (United States)

    Muyama, Hiromi; Kiyokawa, Yasushi; Inagaki, Hideaki; Takeuchi, Yukari; Mori, Yuji

    2016-03-15

    Rats are known to emit a series of ultrasonic vocalizations, termed 22-kHz calls, when exposed to distressing stimuli. Pharmacological studies have indicated that anxiety mediates 22-kHz calls in distressed rats. We previously found that exposure to the rat alarm pheromone increases anxiety in rats. Therefore, we hypothesized that the alarm pheromone would increase 22-kHz calls in pheromone-exposed rats. Accordingly, we tested whether exposure to the alarm pheromone induced 22-kHz calls, as well as whether the alarm pheromone increased 22-kHz calls in response to an aversive conditioned stimulus (CS). Rats were first fear-conditioned to an auditory and contextual CS. On the following day, the rats were either exposed to the alarm pheromone or a control odor that was released from the neck region of odor-donor rats. Then, the rats were re-exposed to the aversive CS. The alarm pheromone neither induced 22-kHz calls nor increased 22-kHz calls in response to the aversive CS. In contrast, the control odor unexpectedly reduced the total number and duration of 22-kHz calls elicited by the aversive CS, as well as the duration of freezing. These results suggest that the alarm pheromone does not affect 22-kHz calls in rats. However, we may have found evidence for an appeasing olfactory signal, released from the neck region of odor-donor rats. Copyright © 2016 Elsevier Inc. All rights reserved.

  7. REACH: impact on the US cosmetics industry?

    Science.gov (United States)

    Pouillot, Anne; Polla, Barbara; Polla, Ada

    2009-03-01

    The Registration, Evaluation, Authorization and restriction of Chemicals (REACH) is a recent European regulation on chemical substances meant to protect human health and the environment. REACH imposes the "precautionary principle" where additional data and definitive action are required when uncertainty is identified. The cosmetics industry is only partially concerned by REACH: while the stages of registration and evaluation apply to cosmetics, those of authorization and restriction most likely will not, as cosmetic ingredients are already subject to regulation by various agencies and directives. REACH has potential benefits to the industry including the possibility of reassuring consumers and improving their image of chemicals and cosmetics. However, REACH also has potential disadvantages, mainly with regard to impeding innovation. The American cosmetics industry will be affected by REACH, because all US manufacturers who export substances to Europe will have to fully comply with REACH.

  8. 40 CFR 267.34 - When must personnel have access to communication equipment or an alarm system?

    Science.gov (United States)

    2010-07-01

    ... to an internal alarm or emergency communication device, either directly or through visual or voice... communication equipment or an alarm system? 267.34 Section 267.34 Protection of Environment ENVIRONMENTAL... have access to communication equipment or an alarm system? (a) Whenever hazardous waste is being poured...

  9. A basic design of alarm system for the future nuclear power plants in Korea

    International Nuclear Information System (INIS)

    Lee, Cheol-Kwon; Hur, Seop; Shin, Jae-Hwal; Koo, In-Soo; Park, Jong-Kyun

    1997-01-01

    The design of an advanced alarm system is under way to apply to the new MMIS for the future nuclear power plants in Korea. Based on the alarm system design bases we established the design requirements and are now refining them with the results of evaluation through the prototype. To realize the advanced system new algorithms for alarm processing and display are implemented and various new devices are examined. The evaluation for the design is performed in accordance with the verification and validation plans and through the prototype. (author). 7 refs, 2 figs

  10. Alarm analysis of secondary loop system based on MFM and SDG methods

    International Nuclear Information System (INIS)

    Yang Ning; Lu Gubing; Chen Pan

    2014-01-01

    The multilevel flow model (MFM) and the signed directed graph (SDG) were combined to analyze the alarm signals of the secondary loop system for nuclear power plant. The MFM was used to delaminate and describe nuclear power plant, and the SDG was used to analyze the logicality of the facility sign in the MFM. Two kinds of faults in the secondary loop system in nuclear power plant were simulated and the alarm signals were analyzed. The simulation results show that the fault source can be identified exactly and the transmit route of the alarm signals can be described clearly, which is helpful for operators to judge. (authors)

  11. Methodology for setting the reference levels in the measurements of the dose rate absorbed in air due to the environmental gamma radiation

    International Nuclear Information System (INIS)

    Dominguez Ley, Orlando; Capote Ferrera, Eduardo; Caveda Ramos, Celia; Alonso Abad, Dolores

    2008-01-01

    Full text: The methodology for setting the reference levels of the measurements of the gamma dose rate absorbed in the air is described. The registration level was obtained using statistical methods. To set the alarm levels, it was necessary to begin with certain affectation level, which activates the investigation operation mode when being reached. It is was necessary to transform this affectation level into values of the indicators selected to set the appearance of an alarm in the network, allowing its direct comparison and at the same time a bigger operability of this one. The affectation level was assumed as an effective dose of 1 mSv/y, which is the international dose limit for public. The conversion factor obtained in a practical way as a consequence of the Chernobyl accident was assumed, converting the value of annual effective dose into values of effective dose rate in air. These factors are the most important in our work, since the main task of the National Network of Environmental Radiological Surveillance of the Republic of Cuba is detecting accidents with a situations regional affectation, and this accident is precisely an example of pollution at this scale. The alarm level setting was based on the results obtained in the first year of the Chernobyl accident. For this purpose, some transformations were achieved. In the final results, a correction factor was introduced depending on the year season the measurement was made. It was taken into account the influence of different meteorological events on the measurement of this indicator. (author)

  12. Functional reach and lateral reach tests adapted for aquatic physical therapy

    Directory of Open Access Journals (Sweden)

    Ana Angélica Ribeiro de Lima

    Full Text Available Abstract Introduction: Functional reach (FR and lateral reach (LR tests are widely used in scientific research and clinical practice. Assessment tools are useful in assessing subjects with greater accuracy and are usually adapted according to the limitations of each condition. Objective: To adapt FR and LR tests for use in an aquatic environment and assess the performance of healthy young adults. Methods: We collected anthropometric data and information on whether the participant exercised regularly or not. The FR and LR tests were adapted for use in an aquatic environment and administered to 47 healthy subjects aged 20-30 years. Each test was repeated three times. Results: Forty-one females and six males were assessed. The mean FR test score for men was 24.06 cm, whereas the mean value for right lateral reach (RLR was 10.94 cm and for left lateral reach (LLR was 9.78 cm. For females, the mean FR score was 17.57 cm, while the mean values for RLR was 8.84cm and for LLR was 7.76 cm. Men performed better in the FR (p < 0.001 and RLR tests than women (p = 0.037. Individuals who exercised regularly showed no differences in performance level when compared with their counterparts. Conclusion: The FR and LR tests were adapted for use in an aquatic environment. Males performed better on the FR and RLR tests, when compared to females. There was no correlation between the FR and LR tests and weight, height, Body Mass Index (BMI, foot length or length of the dominant upper limb.

  13. Real-time Alarm Monitoring System for Detecting Driver Fatigue in Wireless Areas

    Directory of Open Access Journals (Sweden)

    Rongrong Fu

    2017-04-01

    Full Text Available The purpose of this paper was to develop a real-time alarm monitoring system that can detect the fatigue driving state through wireless communication. The drivers’ electroencephalogram (EEG signals were recorded from occipital electrodes. Seven EEG rhythms with different frequency bands as gamma, hbeta, beta, sigma, alpha, theta and delta waves were extracted. They were simultaneously assessed using relative operating characteristic (ROC curves and grey relational analysis to select one as the fatigue feature. The research results showed that the performance of theta wave was the best one. Therefore, theta wave was used as fatigue feature in the following alarm device. The real-time alarm monitoring system based on the result has been developed, once the threshold was settled by using the data of the first ten minutes driving period. The developed system can detect driver fatigue and give alarm to indicate the onset of fatigue automatically.

  14. 30 CFR 75.352 - Actions in response to AMS malfunction, alert, or alarm signals.

    Science.gov (United States)

    2010-07-01

    ..., DEPARTMENT OF LABOR COAL MINE SAFETY AND HEALTH MANDATORY SAFETY STANDARDS-UNDERGROUND COAL MINES Ventilation..., or alarm signal is received at the designated surface location, the sensor(s) that are activated must...

  15. Intelligent buildings, automatic fire alarm and fire-protection control system

    International Nuclear Information System (INIS)

    Tian Deyuan

    1999-01-01

    The author describes in brief the intelligent buildings, and the automatic fire alarm and fire-protection control system. On the basis of the four-bus, three-bus and two-bus, a new transfer technique was developed

  16. COUNTERMEASURE FOR MINIMIZE UNWANTED ALARM OF AUTOMATIC FIRE NOTIFICATION SYSTEM IN THE REPUBLIC OF KOREA

    Directory of Open Access Journals (Sweden)

    Hasung Kong

    2015-01-01

    Full Text Available In this article investigated the cause of error through survey to building officials for minimizing the unwanted alarm of automatic fire notification and suggested countermeasure for minimizing the unwanted alarm. The main cause of the unwanted alarm is defective fire detector, interlocking with automatic fire detection system, lack in fire safety warden’s ability, worn-out fire detect receiving system. The countermeasure for minimizing unwanted alarm is firstly, tightening up the standard of model approval, Secondly, interlocking with cross-section circuit method fire extinguishing system or realizing automatic fire notification system interlocking with home network, thirdly, tightening up licensing examination of fire safety warden, lastly, it suggested term of use rule of fire detect receiving system. 

  17. COUNTERMEASURE FOR MINIMIZE UNWANTED ALARM OF AUTOMATIC FIRE NOTIFICATION SYSTEM IN THE REPUBLIC OF KOREA

    Directory of Open Access Journals (Sweden)

    Hasung Kong

    2015-01-01

    Full Text Available In this article investigated the cause of error through survey to building officials for minimizing the unwanted alarm of automatic fire notification and suggested countermeasure for minimizing the unwanted alarm. The main cause of the unwanted alarm is defective fire detector, interlocking with automatic fire detection system, lack in fire safety warden’s ability, worn-out fire detect receiving system. The countermeasure for minimizing unwanted alarm is firstly, tightening up the standard of model approval, Secondly, interlocking with cross-section circuit method fire extinguishing system or realizing automatic fire notification system interlocking with home network, thirdly, tightening up licensing examination of fire safety warden, lastly, it suggested term of use rule of fire detect receiving system.

  18. Plutonium Finishing Plant (PFP) Criticality Alarm System Commercial Grade Item (CGI) Critical Characteristics

    International Nuclear Information System (INIS)

    WHITE, W.F.

    1999-01-01

    This document specifies the critical characteristics for Commercial Grade Items (CGI) procured for PFP's criticality alarm system as required by HNF-PRO-268 and HNF-PRO-1819. These are the minimum specifications that the equipment must meet in order to properly perform its safety function. There may be several manufacturers or models that meet the critical characteristics for any one item. PFP's Criticality Alarm System includes the nine criticality alarm system panels and their associated hardware. This includes all parts up to the first breaker in the electrical distribution system. Specific system boundaries and justifications are contained in HNF-SD-CP-SDD-003, ''Definition and Means of Maintaining the Criticality Detectors and Alarms Portion of the PFP Safety Envelope.'' The procurement requirements associated with the system necessitates procurement of some system equipment as Commercial Grade Items in accordance with HNF-PRO-268, ''Control of Purchased Items and Services.''

  19. Advances in software development for intelligent interfaces for alarm and emergency management consoles

    International Nuclear Information System (INIS)

    Moseley, M.R.; Olson, C.E.

    1986-01-01

    Recent advances in technology allow features like voice synthesis, voice and speech recognition, image understanding, and intelligent data base management to be incorporated in computer driven alarm and emergency management information systems. New software development environments make it possible to do rapid prototyping of custom applications. Three examples using these technologies are discussed. 1) Maximum use is made of high-speed graphics and voice synthesis to implement a state-of-the-art alarm processing and display system with features that make the operator-machine interface efficient and accurate. 2) An application generator which has the capability of ''building'' a specific alarm processing and display application in a matter of a few hours, using the site definition developed in the security planning phase to produce the custom application. 3) A software tool, is described which permits rapid prototyping of human-machine interfaces for a variety of applications including emergency management, alarm display and process information display

  20. Development of alarm cause tracking system for Korea standard nuclear power plant

    International Nuclear Information System (INIS)

    Lee, Jung Woon; Kim, Jung Taek; Park, Jae Chang; Lee, Hyun Chul; Park, Joong Pal

    2004-05-01

    The proposed system, the ACTS(Alarm Cause Tracking System), in the 1st and 2nd development period(2001. 7 ∼ 2003. 6), tracks and displays the causes of alarms on-line from computerized logic diagrams. And the system highlights the specific procedures related the causes in the procedure of the alarm. In this period(2003. 7 ∼ 2004. 4), we developed the ACTS for Korea standard nuclear power plant. Also, we computerized control logic diagrams and alarm procedures for the ACTS. A long-term target is to apply the ACTS at the real power plant, and a short-term target is to connect the ACTS with the ITF(Intergrated Test Facility) in KAERI site to develop other applications

  1. Alarm symptoms of upper gastrointestinal cancer and contact to general practice

    DEFF Research Database (Denmark)

    Rasmussen, Sanne; Larsen, Pia Veldt; Svendsen, Rikke Pilsgaard

    2015-01-01

    INTRODUCTION: Survival of upper gastrointestinal (GI) cancer depends on early stage diagnosis. Symptom-based guidelines and fast-track referral systems have been implemented for use in general practice. To improve diagnosis of upper GI cancer, knowledge on prevalence of alarm symptoms...... between 1.1% ("repeated vomiting") and 3.4% ("difficulty swallowing"). Women had higher odds of experiencing "repeated vomiting" and "persistent and recent-onset abdominal pain", but lower odds of experiencing "upper GI bleeding". The proportion of people contacting their GP with each of the four specific...... alarm symptoms ranged from 24.3% ("upper GI bleeding") to 39.9% ("repeated vomiting"). For each combination of two specific alarm symptoms, at least 52% contacted their GP. CONCLUSION: The specific alarm symptoms of upper GI cancer are not very prevalent in the general population. The proportion of GP...

  2. Cry features reflect pain intensity in term newborns: an alarm threshold.

    Science.gov (United States)

    Bellieni, Carlo V; Sisto, Renata; Cordelli, Duccio M; Buonocore, Giuseppe

    2004-01-01

    The purpose of this study was to assess differences in sound spectra of crying of term newborns in relation to different pain levels. Fifty-seven consecutively born neonates were evaluated during heel-prick performed with different analgesic techniques. Crying was recorded and frequency spectrograms analyzed. A pain score on the DAN (Douleur Aiguë du Nouveau-né) scale was assigned to each baby after the sampling. Three features were considered and correlated with the corresponding DAN scores: 1) whole spectral form; 2) the fundamental frequency of the first cry emitted (F0); and 3) root mean square sound pressure normalized to its maximum. After emission of the first cry, babies with DAN scores >8, but not with DAN scores cry") characterized by a sequence of almost identical cries with a period on the order of 1 s. A statistically significant correlation was found between root mean square (r2 = 89%, p cry (r2 = 68.2%, p = 0.02), and DAN score. F0 did not show significant correlation with DAN score in the subset of neonates with DAN scores babies with a DAN score >8 had a significantly higher F0 than those with lower DAN scores (p = 0.016). An alarm threshold exists between high (>8) and low (cry at a high pitch is emitted, followed by the siren cry, with a sound level maintained near its maximum.

  3. Cutaneous factitia in elderly patients: alarm signal for psychiatric disorders

    Directory of Open Access Journals (Sweden)

    Chiriac A

    2014-03-01

    choose the proper therapy is mandatory for all these cases. Dermatologists and all physicians who take care of old patients must recognize the disorder in order to provide optimum care for this chronic condition. We emphasize therefore the importance of psychiatric evaluation and treatment to avoid the major risk of suicide. Skin lesions must be regarded as an alarm signal in critical cases, especially in senior people. Keywords: pathomimia, elderly, psychiatric disorders

  4. Dosimetric properties of the pocket alarm dosimeter type Graetz TDW 10/20

    International Nuclear Information System (INIS)

    Hauser, M.; Burgkhardt, B.; Piesch, E.

    1980-08-01

    In personnel monitoring pocket dosimeters with build-in alarm devices are increasingly in use. The report presents results of a test performed at Karlsruhe. The properties investigated are above all linearity and reproducibility of the dose reading as well as of the acustic alarm indication, dependence on the dose reading, the photon energy, the direction of the radiation incidence, the dose rate, the temperature and operational characteristic of the batteries. (orig./HP) [de

  5. Integrated knowledge base tool for acquisition and verification of NPP alarm systems

    International Nuclear Information System (INIS)

    Park, Joo Hyun; Seong, Poong Hyun

    1998-01-01

    Knowledge acquisition and knowledge base verification are important activities in developing knowledge-based systems such as alarm processing systems. In this work, we developed the integrated tool, for knowledge acquisition and verification of NPP alarm processing systems, by using G2 tool. The tool integrates document analysis method and ECPN matrix analysis method, for knowledge acquisition and knowledge verification, respectively. This tool enables knowledge engineers to perform their tasks from knowledge acquisition to knowledge verification consistently

  6. Game Theoretic Approach for Systematic Feature Selection; Application in False Alarm Detection in Intensive Care Units

    Directory of Open Access Journals (Sweden)

    Fatemeh Afghah

    2018-03-01

    Full Text Available Intensive Care Units (ICUs are equipped with many sophisticated sensors and monitoring devices to provide the highest quality of care for critically ill patients. However, these devices might generate false alarms that reduce standard of care and result in desensitization of caregivers to alarms. Therefore, reducing the number of false alarms is of great importance. Many approaches such as signal processing and machine learning, and designing more accurate sensors have been developed for this purpose. However, the significant intrinsic correlation among the extracted features from different sensors has been mostly overlooked. A majority of current data mining techniques fail to capture such correlation among the collected signals from different sensors that limits their alarm recognition capabilities. Here, we propose a novel information-theoretic predictive modeling technique based on the idea of coalition game theory to enhance the accuracy of false alarm detection in ICUs by accounting for the synergistic power of signal attributes in the feature selection stage. This approach brings together techniques from information theory and game theory to account for inter-features mutual information in determining the most correlated predictors with respect to false alarm by calculating Banzhaf power of each feature. The numerical results show that the proposed method can enhance classification accuracy and improve the area under the ROC (receiver operating characteristic curve compared to other feature selection techniques, when integrated in classifiers such as Bayes-Net that consider inter-features dependencies.

  7. REACH: Evaluation Report and Executive Summary

    Science.gov (United States)

    Sibieta, Luke

    2016-01-01

    REACH is a targeted reading support programme designed to improve reading accuracy and comprehension in pupils with reading difficulties in Years 7 and 8. It is based on research by the Centre for Reading and Language at York and is delivered by specially trained teaching assistants (TAs). This evaluation tested two REACH interventions, one based…

  8. Alarm systems detect volcanic tremor and earthquake swarms during Redoubt eruption, 2009

    Science.gov (United States)

    Thompson, G.; West, M. E.

    2009-12-01

    We ran two alarm algorithms on real-time data from Redoubt volcano during the 2009 crisis. The first algorithm was designed to detect escalations in continuous seismicity (tremor). This is implemented within an application called IceWeb which computes reduced displacement, and produces plots of reduced displacement and spectrograms linked to the Alaska Volcano Observatory internal webpage every 10 minutes. Reduced displacement is a measure of the amplitude of volcanic tremor, and is computed by applying a geometrical spreading correction to a displacement seismogram. When the reduced displacement at multiple stations exceeds pre-defined thresholds and there has been a factor of 3 increase in reduced displacement over the previous hour, a tremor alarm is declared. The second algorithm was to designed to detect earthquake swarms. The mean and median event rates are computed every 5 minutes based on the last hour of data from a real-time event catalog. By comparing these with thresholds, three swarm alarm conditions can be declared: a new swarm, an escalation in a swarm, and the end of a swarm. The end of swarm alarm is important as it may mark a transition from swarm to continuous tremor. Alarms from both systems were dispatched using a generic alarm management system which implements a call-down list, allowing observatory scientists to be called in sequence until someone acknowledged the alarm via a confirmation web page. The results of this simple approach are encouraging. The tremor alarm algorithm detected 26 of the 27 explosive eruptions that occurred from 23 March - 4 April. The swarm alarm algorithm detected all five of the main volcanic earthquake swarm episodes which occurred during the Redoubt crisis on 26-27 February, 21-23 March, 26 March, 2-4 April and 3-7 May. The end-of-swarm alarms on 23 March and 4 April were particularly helpful as they were caused by transitions from swarm to tremor shortly preceding explosive eruptions; transitions which were

  9. Sound the alarm : monitoring system's real-time data reduces well downtime

    Energy Technology Data Exchange (ETDEWEB)

    Harrison, L.

    2008-10-15

    This article presented a new electrical heating element tool used to boost performance at low-producing wells. The down hole tool (DHT) was designed to generate steam underground in order to avoid the energy losses associated with above-ground equipment. The tool operates like an oven element but is connected to a computerized well-monitoring system in order to ensure that heat is evenly distributed. The DHT is equipped with an Optiview monitoring system designed to provide continuous readouts of near real time data. The system can also be customized to deliver messages to telephones, computers, or alarm systems. Changes in hydraulic torque, flow rates, tank levels, speeds, temperatures and hydraulic pressures can be set within ranges in order to optimize oil and gas well production by reducing operating costs and increasing equipment efficiencies. Use of the tool has during tests has minimized the need for well servicing. The patent pending technology is also being investigated for its use in water disposal and other pump-to-surface units. It was concluded that use of the technology will allow oil and gas managers to determine the status of wells in remote locations. 1 fig.

  10. Radiation alarms and access control systems: Recommendations of the National Council on Radiation Protection and Measurements

    International Nuclear Information System (INIS)

    1986-01-01

    In facilities where radioactive materials are handled, or where radiation-producing equipment is used, the building, the equipment, and the associated safety procedures should be designed and developed together to provide a safe work environment. The specific combination of requirements for a given facility is defined by the operational radiation safety program. It should be emphasized that this report describes a range of alarm and access control systems that can provide an acceptable level of safety at many types of facilities. Depending on circumstances, the solutions offered may not be appropriate for certain facilities because they are to restrictive, not restrictive enough, or do not cover all circumstances. The document is offered as a starting point providing ideas that professional health physicists can adapt to meet the needs of a particular situation. Under no circumstances should this report be interpreted in ''cookbook'' fashion, with literal adherence to every recommendation demanded, nor should it be expected to provide adequate protection in every case without consideration of local conditions

  11. The database for reaching experiments and models.

    Directory of Open Access Journals (Sweden)

    Ben Walker

    Full Text Available Reaching is one of the central experimental paradigms in the field of motor control, and many computational models of reaching have been published. While most of these models try to explain subject data (such as movement kinematics, reaching performance, forces, etc. from only a single experiment, distinct experiments often share experimental conditions and record similar kinematics. This suggests that reaching models could be applied to (and falsified by multiple experiments. However, using multiple datasets is difficult because experimental data formats vary widely. Standardizing data formats promises to enable scientists to test model predictions against many experiments and to compare experimental results across labs. Here we report on the development of a new resource available to scientists: a database of reaching called the Database for Reaching Experiments And Models (DREAM. DREAM collects both experimental datasets and models and facilitates their comparison by standardizing formats. The DREAM project promises to be useful for experimentalists who want to understand how their data relates to models, for modelers who want to test their theories, and for educators who want to help students better understand reaching experiments, models, and data analysis.

  12. Communication strategies and timeliness of response to life critical telemetry alarms.

    Science.gov (United States)

    Bonzheim, Kimberly A; Gebara, Rani I; O'Hare, Bridget M; Ellis, R Darin; Brand, Monique A; Balar, Salil D; Stockman, Rita; Sciberras, Annette M; Haines, David E

    2011-05-01

    A centralized electrocardiogram telemetry monitoring system (TMS) facilitates early identification of critical arrhythmias and acute medical decompensation. Timely intervention can only be performed if abnormalities are communicated rapidly to the direct caregiver. The study objectives were to measure effectiveness of bi-directional voice communication badges versus one-way alphanumeric pagers for telemetry alarm response and communication loop closure. A sequential observational pilot study of nursing response to TMS alarms compared communication technologies on four nursing units in a 1,061 bed tertiary care hospital with 264 TMS channels of telemetry over a 2-year period. Subsequently, the communication technologies were compared in a randomized fashion on a 68-bed progressive cardiac care unit. Caregivers were blinded to the protocol. All alarm responses were recorded during two periods using either pagers or voice communication devices. Alarm response time and closure of the communication loop were analyzed in a blinded fashion. The direct communication functionality of the badge significantly shortened the time to first contact, time to completion, and rate of closure of the communication loop in both the pilot and study phases. Median time to first contact with the communication badge was 0.5  min, compared to 1.6  min with pager communication (p Communication loop closure was achieved in 100% of clinical alarms using the badge versus 19% with the pager (p Communication badge technology reduced alarm time to first contact and completion as well as facilitated communication loop closures. Immediate two-way communication significantly impacted practice, alarm management, and resulted in faster bedside care.

  13. A new diagnosis method using alarm annunciation for FBR power plants

    International Nuclear Information System (INIS)

    Ozaki, Y.; Suda, K.; Yoshikawa, S.; Ozawa, K.

    1997-01-01

    We discuss the methodology diversity for diagnosis reasoning in autonomous operation system, and propose a new diagnosis method using alarm annunciation system. The methodology diversity is assured by preparing plural agents, each of which is based on its own different methodology, therefore, it is expected for the reliability in diagnosis to be improved. Meanwhile, the combination of annunciated alarms is expected to be peculiar to the anomalous phenomenon or accident. Moreover, as the state of affairs is developing, each appearance of the pattern is changing with time peculiarly to each anomaly or accident. The matter is utilized for the new diagnosis method. The patterns of annunciated alarms with progress of the events are prepared in advance under the condition of the anomalies or accidents by use of plant simulatory. The diagnostic reasoning can be done by comparing the obtained combination of annunciated alarms with the reference templates, pattern matching method. On the other hand, we have another method, called as COBWEB used for conceptual classification in cognitive science, to reason for diagnosis. We have carried out the experiments using the loop type LMFBR plant simulator to obtain the various combinations of annunciated alarms with progress of the events under the conditions of anomalies and accidents. The examined cases were related to the anomalies and accidents in the water/steam system of the LMFBR power plant. We have obtained the conclusions that it is effective to reason the causes of anomalies using the annunciated alarms. We are going to apply the pattern matching technique or COBWEB method into the diagnostic reasoning to confirm the performance of the proposed diagnosis method based on the alarm annunciation. (author). 5 refs, 14 figs

  14. A new diagnosis method using alarm annunciation for nuclear power plants

    International Nuclear Information System (INIS)

    Ozaki, Yoshihiko; Suda, Kazunori; Ozawa, Kenji

    1997-01-01

    We discuss the methodology diversity for diagnosis reasoning in an autonomous operation system, and propose a new diagnosis method using an alarm annunciation system. The combination of annunciated alarms is expected to be peculiar to the anomalous phenomenon or accident. Moreover, as the state of affairs is developing, each appearance of the pattern is changing with time peculiarly to each anomaly or accident. The matter is utilized for the new diagnosis method. The patterns of annunciated alarms with progress of the events are prepared in advance under the condition of the anomalies or accidents by use of a plant simulator. The diagnostic reasoning can be done by comparing the obtained combination of annunciated alarms with the reference templates by using pattern matching method. On the other hand, we have another method, called COBWEB used for conceptual classification in cognitive science, to reason for diagnosis. We have carried out the experiments using the loop type LMFBR plant simulator to obtain the various combinations of annunciated alarms with progress of the events under the conditions of anomalies and accidents. The examined cases were related to the anomalies and accidents in the water/steam system of the LMFBR power plant. The simulation examination showed that each change of the pattern of annunciated alarms is specific to each anomaly or accident, and we have applied the pattern matching technique and COBWEB methods into the diagnostic reasoning using annunciated alarms. We could show the capability of these two methods to reason and focus among various candidates of causes of anomalies with gradually improved conviction degree as time passes from the occurrence of anomalies. It was also confirmed that these methods are effective in diagnosis reasoning as a way the operators are doing the diagnosis reasoning in existing plants. (author)

  15. The acute physiological stress response to an emergency alarm and mobilization during the day and at night.

    Science.gov (United States)

    Hall, Sarah J; Aisbett, Brad; Tait, Jamie L; Turner, Anne I; Ferguson, Sally A; Main, Luana C

    2016-01-01

    The purpose of this study was to investigate the acute physiological stress response to an emergency alarm and mobilization during the day and at night. Sixteen healthy males aged 25 ± 4 years (mean ± SD) spent four consecutive days and nights in a sleep laboratory. This research used a within-participants design with repeated measures for time, alarm condition (alarm or control), and trial (day or night). When an alarm sounded, participants were required to mobilize immediately. Saliva samples for cortisol analysis were collected 0 min, 15 min, 30 min, 45 min, 60 min, 90 min, and 120 min after mobilization, and at corresponding times in control conditions. Heart rate was measured continuously throughout the study. Heart rate was higher in the day (F(20,442) = 9.140, P night (F(23,459) = 8.356, P day alarm and day control conditions. Cortisol was higher (F(6,183) = 2.450, P night alarm and mobilization compared to the night control condition. The magnitude of difference in cortisol between night control and night alarm conditions was greater (F(6,174) = 4.071, P day control and day alarm conditions. The augmented heart rate response to the day and night alarms supports previous observations in field settings. Variations in the cortisol responses between conditions across the day and night may relate to differences in participants' ability to interpret the alarm when sleeping versus when awake.

  16. Alarming Levels of Drug-Resistant Tuberculosis in HIV-Infected Patients in Metropolitan Mumbai, India

    OpenAIRE

    Isaakidis, Petros; Das, Mrinalini; Kumar, Ajay M V; Peskett, Christopher; Khetarpal, Minni; Bamne, Arun; Adsul, Balkrishna; Manglani, Mamta; Sachdeva, Kuldeep Singh; Parmar, Malik; Kanchar, Avinash; Rewari, B B; Deshpande, Alaka; Rodrigues, Camilla; Shetty, Anjali

    2014-01-01

    BACKGROUND: Drug-resistant tuberculosis (DR-TB) is a looming threat to tuberculosis control in India. However, no countrywide prevalence data are available. The burden of DR-TB in HIV-co-infected patients is likewise unknown. Undiagnosed and untreated DR-TB among HIV-infected patients is a major cause of mortality and morbidity. We aimed to assess the prevalence of DR-TB (defined as resistance to any anti-TB drug) in patients attending public antiretroviral treatment (ART) centers in greater ...

  17. Implementation guidance for industrial-level security systems using radio frequency alarm links

    Energy Technology Data Exchange (ETDEWEB)

    Swank, R.G.

    1996-07-12

    Spread spectrum (SS) RF transmission technologies have properties that make the transmitted signal difficult to intercept, interpret, and jam. The digital code used in the modulation process results in a signal that has high reception reliability and supports multiple use of frequency bands and selective addressing. These attributes and the relatively low installation cost of RF systems make SSRF technologies candidate for communications links in security systems used for industrial sites, remote locations, and where trenching or other disturbances of soil or structures may not be desirable or may be costly. This guide provides a description of such a system and presents implementation methods that may be of engineering benefit.

  18. Enhancing US Operational Reach in Southeast Asia

    National Research Council Canada - National Science Library

    Hitchcock, David

    2003-01-01

    .... While this treat continues to exist, the US Pacific Command (PACOM) must also pursue a neat term methodology to expand its operational reach and ability to respond to contingencies throughout the East Asian littoral, especially within Southeast Asia...

  19. Reaching the Overlooked Student in Physical Education

    Science.gov (United States)

    Esslinger, Keri; Esslinger, Travis; Bagshaw, Jarad

    2015-01-01

    This article describes the use of live action role-playing, or "LARPing," as a non-traditional activity that has the potential to reach students who are not interested in traditional physical education.

  20. Compact muon solenoid magnet reaches full field

    CERN Multimedia

    2006-01-01

    Scientist of the U.S. Department of Energy in Fermilab and collaborators of the US/CMS project announced that the world's largest superconducting solenoid magnet has reached full field in tests at CERN. (1 apge)

  1. A fence with several safety wires with alarm equipment

    International Nuclear Information System (INIS)

    Osterburg, G.

    1977-01-01

    The patent claim concerns a safety arrangement for objects in the open against unauthorised interference. A fence provided with several safety wires is used for this purpose. Each wire has a piezo-electric transmitter at one end, and has a piezo-electric receiver at the other end, which converts the mechanical vibrations into electrical signals. The piezo-electric receiver is connected via an amplifier to the piezo-electric transmitter of the next wire, which initiates the latter to produce mechanical vibrations and so on, until the transmitter of the first wire is reached again, so that a closed feed-back loop is formed. An electronic evaluation circuit is connected between one of the amplfiers and the subsequent transmitter, which on loss of vibration due to touching, hold. (ORU) [de

  2. When Does the Warmest Water Reach Greenland?

    Science.gov (United States)

    Grist, J. P.; Josey, S. A.; Boehme, L.; Meredith, M. P.; Laidre, K. L.; Heide-Jørgensen, M. P.; Kovacs, K. M.; Lydersen, C.; Davidson, F. J. M.; Stenson, G. B.; Hammill, M. O.; Marsh, R.; Coward, A.

    2016-02-01

    The warmest water reaching the east and west coast of Greenland is found between 200 and 600 m, in the warm Atlantic Water Layer (WL). Temperature changes within the WL have been highlighted as a possible cause of accelerated melting of tidewater glaciers and therefore are an important consideration for understanding global sea level rise. However, a limited number of winter observations of the WL have prohibited determining its seasonal variability. To address this, temperature data from Argo profiling floats, a range of sources within the World Ocean Database, and unprecedented coverage from marine-mammal borne sensors have been analyzed for the period 2002-2011. A significant seasonal range in temperature ( 1-2°C) is found in the warm layer, in contrast to most of the surrounding ocean. The magnitude of the seasonal cycle is thus comparable with the 1990s warming that was associated with an increased melt rate in a marine terminating glacier of West Greenland. The phase of the seasonal cycle exhibits considerable spatial variability; with high-resolution ocean model trajectory analysis suggesting it is determined by the time taken for waters to be advected from the subduction site in the Irminger Basin. For western Greenland, the annual temperature maximum occurs near or after the turn of the calendar year. This is significant because a recent study suggested that it is in the non-summer months when fjord-shelf exchanges allow the WL to most strongly influence glacier melt rate. However this is also the time of the year when the WL is least well observed. It is therefore clear that year-round subsurface temperature measurements are still required for a complete description of the WL seasonality, and in particular to ensure that the ice-melting potential of the WL is not underestimated.

  3. Development of alarm logics for critical function monitoring in SMART-P reactor

    Energy Technology Data Exchange (ETDEWEB)

    Seong, Seung Hwan; Hur, Seop; Seo, Jae Kwang; Lee, Tae Ho; Park, Cheon Tae; Kang, Han Ok

    2003-04-01

    The alarm logics for the critical functions of SMART-P reactor are developed, which are based on the those of Korean Standard Nuclear power Plant(KSNP). The SMART-P reactor is an integral typed nuclear power plant and has the some different design features compared to those of KSNP. It, however, has the similar features in critical functions because it is a kind of pressurized water reactor. The alarm logics for Critical Function Monitoring System(CFMS) in SMART-P are developed from those for CFMS in KSNP. The alarm logics of CFMS in only the primary loop are, therefore, developed, though the general CFMS covered the status of primary and secondary loop including the features of the containment. The specific setpoint of related variables related to the alarm logics will be developed after the specific designs of SMART-P are finished. In appendix, we describe the conceptual architecture and variables of display screens on CFMS according to the developed alarm logics.

  4. Benefit of an electronic medical record-based alarm in the optimization of stress ulcer prophylaxis.

    Science.gov (United States)

    Saad, Emanuel José; Bedini, Marianela; Becerra, Ana Florencia; Martini, Gustavo Daniel; Gonzalez, Jacqueline Griselda; Bolomo, Andrea; Castellani, Luciana; Quiroga, Silvana; Morales, Cristian; Leathers, James; Balderramo, Domingo; Albertini, Ricardo Arturo

    2018-06-09

    The use of stress ulcer prophylaxis (SUP) has risen in recent years, even in patients without a clear indication for therapy. To evaluate the efficacy of an electronic medical record (EMR)-based alarm to improve appropriate SUP use in hospitalized patients. We conducted an uncontrolled before-after study comparing SUP prescription in intensive care unit (ICU) patients and non-ICU patients, before and after the implementation of an EMR-based alarm that provided the correct indications for SUP. 1627 patients in the pre-intervention and 1513 patients in the post-intervention cohorts were included. The EMR-based alarm improved appropriate (49.6% vs. 66.6%, p<0.001) and reduced inappropriate SUP use (50.4% vs. 33.3%, p<0.001) in ICU patients only. These differences were related to the optimization of SUP in low risk patients. There was no difference in overt gastrointestinal bleeding between the two cohorts. Unjustified costs related to SUP were reduced by a third after EMR-based alarm use. The use of an EMR-based alarm improved appropriate and reduced inappropriate use of SUP in ICU patients. This benefit was limited to optimization in low risk patients and associated with a decrease in SUP costs. Copyright © 2018 Elsevier España, S.L.U. All rights reserved.

  5. "Alarm-corrected" ergonomic armrest use could improve learning curves of novices on robotic simulator.

    Science.gov (United States)

    Yang, Kun; Perez, Manuela; Hossu, Gabriela; Hubert, Nicolas; Perrenot, Cyril; Hubert, Jacques

    2017-01-01

    In robotic surgery, the professional ergonomic habit of using an armrest reduces operator fatigue and increases the precision of motion. We designed and validated a pressure surveillance system (PSS) based on force sensors to investigate armrest use. The objective was to evaluate whether adding an alarm to the PSS system could shorten ergonomic training and improve performance. Twenty robot and simulator-naïve participants were recruited and randomized in two groups (A and B). The PSS was installed on a robotic simulator, the dV-Trainer, to detect contact with the armrest. The Group A members completed three tasks on the dV-Trainer without the alarm, making 15 attempts at each task. The Group B members practiced the first two tasks with the alarm and then completed the final tasks without the alarm. The simulator provided an overall score reflecting the trainees' performance. We used the new concept of an "armrest load" score to describe the ergonomic habit of using the armrest. Group B had a significantly higher performance score (p ergonomic errors and accelerated professional ergonomic habit acquisition. The combination of the PSS and alarm is effective in significantly shortening the learning curve in the robotic training process.

  6. Advances in software development for intelligent interfaces for alarm and emergency management consoles

    International Nuclear Information System (INIS)

    Moseley, M.R.; Olson, C.E.

    1986-01-01

    Recent advances in technology allow features like voice synthesis, voice and speech recognition, image understanding, and intelligent data base management to be incorporated in computer driven alarm and emergency management information systems. New software development environments make it possible to do rapid prototyping of custom applications. Three examples using these technologies are discussed. (1) Maximum use is made of high-speed graphics and voice synthesis to implement a state-of-the-art alarm processing and display system with features that make the operator-machine interface efficient and accurate. Although very functional, this system is not portable or flexible; the software would have to be substantially rewritten for other applications. (2) An application generator which has the capability of ''building'' a specific alarm processing and display application in a matter of a few hours, using the site definition developed in the security planning phase to produce the custom application. This package is based on a standardized choice of hardware, within which it is capable of building a system to order, automatically constructing graphics, data tables, alarm prioritization rules, and interfaces to peripherals. (3) A software tool, the User Interface Management System (UIMS), is described which permits rapid prototyping of human-machine interfaces for a variety of applications including emergency management, alarm display and process information display. The object-oriented software of the UIMS achieves rapid prototyping of a new interface by standardizing to a class library of software objects instead of hardware objects

  7. Reaching the Millennium Development Goals : Mauritania Should Care

    OpenAIRE

    Magnoli Bocchi, Alessandro; Pontara, Nicola; Fall, Khayar; Tejada, Catalina M.; Gallego Cuervo, Pablo

    2008-01-01

    Mauritania is a resource-rich developing country. As many other African nations, it will not reach most of the Millennium Development Goals, unless the authorities commit to accelerating progress. To succeed by 2015, the government needs to: mobilize additional financial resources, introduce policy changes at the sector level, and strengthen the links between strategic objectives and the b...

  8. Reaching remote areas in Latin America.

    Science.gov (United States)

    Jaimes, R

    1994-01-01

    Poor communities in remote and inaccessible areas tend to not only be cut off from family planning education and services, but they are also deprived of basic primary health care services. Efforts to bring family planning to such communities and populations should therefore be linked with other services. The author presents three examples of programs to bring effective family planning services to remote communities in Central and South America. Outside of the municipal center in the Tuxtlas region of Mexico, education and health levels are low and people live according to ancient customs. Ten years ago with the help of MEXFAM, the IPPF affiliate in Mexico, two social promoters established themselves in the town of Catemaco to develop a community program of family planning and health care offering education and prevention to improve the quality of people's lives. Through their health brigades taking health services to towns without an established health center, the program has influenced an estimated 100,000 people in 50 villages and towns. The program also has a clinic. In Guatemala, the Family Welfare Association (APROFAM) gave bicycles to 240 volunteer health care workers to facilitate their outreach work in rural areas. APROFAM since 1988 has operated an integrated program to treat intestinal parasites and promote family planning in San Lucas de Toliman, an Indian town close to Lake Atitlan. Providing health care to more than 10,000 people, the volunteer staff has covered the entire department of Solola, reaching each family in the area. Field educators travel on motorcycles through the rural areas of Guatemala coordinating with the health volunteers the distribution of contraceptives at the community level. The Integrated Project's Clinic was founded in 1992 and currently carries out pregnancy and Pap tests, as well as general lab tests. Finally, Puna is an island in the middle of the Gulf of Guayaquil, Ecuador. Women on the island typically have 10

  9. Improvement and test of alarm cause tracking system for Kori nuclear power plant units 3 and 4

    International Nuclear Information System (INIS)

    Lee, Jung Woon; Park, Jae Chang; Lee, Jung Woon; Kim, Jung Taek; Lyu, Sung Pil; Kim, Eun Ju; Park, Joong Pal

    2003-05-01

    The proposed system, ACTS(Alarm Cause Tracking System), in the 1st development period(2002. 7 ∼ 2003. 6), tracks and displays the causes of alarms on-line from computerized logic diagrams. And the system highlights the specific procedures related the causes in the procedure of the alarm. But, some problems were found in ACTS on editing logic diagram and logic processing in realtime for 2000 logic diagrams. In 2nd development period, we improved the data structure of graphic information for logic diagram and changed function oriented programming to object oriented programming for logic elements. Also, the display of precedent alarms which introduce many following alarms is provided to avoid confusion from the followed nuisance alarms. And logic input signal generator was developed to test the ACTS which generates input signal in time sequence of events acquired from simulator or real plant

  10. Two odorant-binding proteins mediate the behavioural response of aphids to the alarm pheromone (E-ß-farnesene and structural analogues.

    Directory of Open Access Journals (Sweden)

    Yu Feng Sun

    Full Text Available Aphids are agricultural pests of great economical interest. Alternatives to insecticides, using semiochemicals, are of difficult applications. In fact, sex pheromones are of little use as aphids reproduce partenogenetically most of the time. Besides, the alarm pheromone, (E-ß-farnesene for a great number of species, is difficult to synthesize and unstable in the environment. The search for novel semiochemicals to be used in population control can be efficiently approached through the study of the olfactory system at the biochemical level. Recently odorant-binding proteins (OBPs have been shown to play a central role in olfactory recognition, thus becoming the target of choice for designing new semiochemicals.To address the question of how the alarm message is recognised at the level of OBPs, we have tested 29 compounds, including (E-ß-farnesene, in binding assays with 6 recombinant proteins and in behaviour experiments. We have found that good repellents bind OBP3 and/or OBP7, while non repellents present different spectra of binding. These results have been verified with two species of aphids, Acyrthosiphon pisum and Myzus persicae, both using (E-ß-farnesene as the alarm pheromone.Our results represent further support to the idea (so far convincingly demonstrated only in Drosophila that OBPs are involved in decoding the chemical information of odorants and pheromones, and for the first time provide such evidence in other insect species and using wild-type insects. Moreover, the data offer guidelines and protocols for the discovery of potential alarm pheromones, using ligand-binding assays as a preliminary screening before subjecting selected compounds to behaviour tests.

  11. Associations between health care seeking and socioeconomic and demographic determinants among people reporting alarm symptoms of cancer

    DEFF Research Database (Denmark)

    Svendsen, Rikke P; Jarbol, Dorte E; Larsen, Pia V

    2013-01-01

    Late diagnosis of cancer may partly be explained by the fact that some patients do not seek health care promptly when experiencing an alarm symptom. Socioeconomic and demographic differences exist concerning knowledge and awareness of cancer alarm symptoms in the general population...... and socioeconomic differences are found in cancer incidence and survival. We therefore hypothesise that socioeconomic and demographic differences in health care-seeking behaviour are present among people with alarm symptoms....

  12. Sequence and batch language programs and alarm related C Programs for the 242-A MCS

    International Nuclear Information System (INIS)

    Berger, J.F.

    1996-01-01

    A Distributive Process Control system was purchased by Project B-534, 242-A Evaporator/Crystallizer Upgrades. This control system, called the Monitor and Control system (MCS), was installed in the 242-A evaporator located in the 200 East Area. The purpose of the MCS is to monitor and control the Evaporator and monitor a number of alarms and other signals from various Tank Farm facilities. Applications software for the MCS was developed by the Waste Treatment Systems Engineering (WTSE) group of Westinghouse. The standard displays and alarm scheme provide for control and monitoring, but do not directly indicate the signal location or depict the overall process. To do this, WTSE developed a second alarm scheme

  13. Associations between reporting of cancer alarm symptoms and socioeconomic and demographic determinants

    DEFF Research Database (Denmark)

    Svendsen, Rikke Pilsgaard; Paulsen, Maja Skov; Larsen, Pia Veldt

    2012-01-01

    ABSTRACT: BACKGROUND: Reporting of symptoms which may signal cancer is the first step in the diagnostic pathway of cancer diseases. Cancer alarm symptoms are common in the general population. Public awareness and knowledge of cancer symptoms are sparse, however, and many people do not seek medical...... help when having possible cancer symptoms. As social inequality is associated with cancer knowledge, cancer awareness, and information-seeking, our hypothesis is that social inequality may also exist in the general population with respect to reporting of cancer alarm symptoms. The aim of this study...... was to investigate possible associations between socioeconomic and demographic determinants and reporting of common cancer alarm symptoms. METHODS: A cross-sectional questionnaire survey was performed based on a stratified sample of the Danish general population. A total of 13 777 randomly selected persons aged 20...

  14. The study of system function analysis method for success path alarm design

    International Nuclear Information System (INIS)

    Kang, S. K.; Shin, Y. C.

    1999-01-01

    The key benefit to the common use of the critical function approach for safety and mission functions is that monitoring methods expected to be used by operaotrs during emergency condition are used continuously during normal operation. For each critical safety function there exists two or more success paths. Information Processing System monitors the availability, operation state and performance of the critical function success paths. In this paper, We have studied System Function Analysis(SFA) for the design of Success Path Alarm(SPA) for applying in KNGR. In here, we thought that SFA will help the design of SPA. The SFA can be applicable to the design of SPA according to NUREG-0711, also can induce the algorithm for alarm of system, train and flow path. We present a method of system function analysis for designing Success Path Alarm

  15. Do working environment interventions reach shift workers?

    Science.gov (United States)

    Nabe-Nielsen, Kirsten; Jørgensen, Marie Birk; Garde, Anne Helene; Clausen, Thomas

    2016-01-01

    Shift workers are exposed to more physical and psychosocial stressors in the working environment as compared to day workers. Despite the need for targeted prevention, it is likely that workplace interventions less frequently reach shift workers. The aim was therefore to investigate whether the reach of workplace interventions varied between shift workers and day workers and whether such differences could be explained by the quality of leadership exhibited at different times of the day. We used questionnaire data from 5361 female care workers in the Danish eldercare sector. The questions concerned usual working hours, quality of leadership, and self-reported implementation of workplace activities aimed at stress reduction, reorganization of the working hours, and participation in improvements of working procedures or qualifications. Compared with day workers, shift workers were less likely to be reached by workplace interventions. For example, night workers less frequently reported that they had got more flexibility (OR 0.5; 95 % CI 0.3-0.7) or that they had participated in improvements of the working procedures (OR 0.6; 95 % CI 0.5-0.8). Quality of leadership to some extent explained the lack of reach of interventions especially among fixed evening workers. In the light of the evidence of shift workers' stressful working conditions, we suggest that future studies focus on the generalizability of results of the present study and on how to reach this group and meet their needs when designing and implementing workplace interventions.

  16. An alarm filtering system for an automated process: a multiple-agent approach

    International Nuclear Information System (INIS)

    Khoualdi, Kamel

    1994-01-01

    Nowadays, the supervision process of industrial installations is more and more complex involving the automation of their control. A malfunction generates an avalanche of alarms. The operator, in charge of the supervision, must face the incident and execute right actions to recover a normal situation. Generally, he is drowned under the great number of alarms. Our aim, in the frame of our researches, is to perform an alarm filtering system for an automated metro line, to help the operator finding the main alarm responsible for the malfunction. Our works are divided into two parts, both dealing with study and development of an alarm filtering system but using two different approaches. The first part is developed in the frame of the SARA project (an operator assistance system for an automated metro line) which is an expert system prototype helping the operators of a command center. In this part, a centralized approach has been used representing the events with a single event graph and using a global procedure to perform diagnosis. This approach has itself shown its limits. In the second part of our works, we have considered the distributed artificial intelligence (DAI) techniques, and more especially the multi-agent approach. The multi-agent approach has been motivated by the natural distribution of the metro line equipment and by the fact that each equipment has its own local control and knowledge. Thus, each equipment has been considered as an autonomous agent. Through agents cooperation, the system is able to determine the main alarm and the faulty equipment responsible for the incident. A prototype, written in SPIRAL (a tool for knowledge-based system) is running on a workstation. This prototype has allowed the concretization and the validation of our multi-agent approach. (author) [fr

  17. Network-Based Real-time Integrated Fire Detection and Alarm (FDA) System with Building Automation

    Science.gov (United States)

    Anwar, F.; Boby, R. I.; Rashid, M. M.; Alam, M. M.; Shaikh, Z.

    2017-11-01

    Fire alarm systems have become increasingly an important lifesaving technology in many aspects, such as applications to detect, monitor and control any fire hazard. A large sum of money is being spent annually to install and maintain the fire alarm systems in buildings to protect property and lives from the unexpected spread of fire. Several methods are already developed and it is improving on a daily basis to reduce the cost as well as increase quality. An integrated Fire Detection and Alarm (FDA) systems with building automation was studied, to reduce cost and improve their reliability by preventing false alarm. This work proposes an improved framework for FDA system to ensure a robust intelligent network of FDA control panels in real-time. A shortest path algorithmic was chosen for series of buildings connected by fiber optic network. The framework shares information and communicates with each fire alarm panels connected in peer to peer configuration and declare the network state using network address declaration from any building connected in network. The fiber-optic connection was proposed to reduce signal noises, thus increasing large area coverage, real-time communication and long-term safety. Based on this proposed method an experimental setup was designed and a prototype system was developed to validate the performance in practice. Also, the distributed network system was proposed to connect with an optional remote monitoring terminal panel to validate proposed network performance and ensure fire survivability where the information is sequentially transmitted. The proposed FDA system is different from traditional fire alarm and detection system in terms of topology as it manages group of buildings in an optimal and efficient manner.Introduction

  18. Validation of the Rome III criteria and alarm symptoms for recurrent abdominal pain in children.

    Science.gov (United States)

    Gijsbers, Carolien F M; Benninga, Marc A; Schweizer, Joachim J; Kneepkens, C M Frank; Vergouwe, Yvonne; Büller, Hans A

    2014-06-01

    Rome criteria were formulated to define functional gastrointestinal disorders (Rome III criteria, 2006) excluding organic diagnoses when alarm symptoms were absent. The aims of the study were to validate the Rome III criteria as to their capacity to differentiate between organic and functional abdominal pain and to assess the role of alarm symptoms in this differentiation. During 2 years all of the patients (ages 4-16 years) presenting with recurrent abdominal pain (Apley criteria) and referred to secondary care were included. Clinical diagnoses were based on protocolized evaluation and intervention with 6-month follow-up. Alarm symptoms were registered. Rome III criteria for functional pain syndromes were assigned independently. Descriptive statistical analyses were performed. In 200 patients (87 boys, mean age 8.8 years), organic (17%), functional (40%), combined organic and functional (9%), spontaneous recovery (27%), and other (8%) clinical diagnoses were established. Alarm symptoms were found in 57.5% (organic causes 56%, functional causes 61%). The evaluation for Rome symptom clusters revealed symptoms of irritable bowel syndrome in 27%, functional dyspepsia in 15%, functional abdominal pain in 28%, functional abdominal pain syndrome in 14.5%, and no pain syndrome in 15.5%. Rome diagnoses, based on symptoms and absence of alarm symptoms, predicted functional clinical diagnosis with sensitivity 0.35 (95% confidence interval 0.27-0.43), specificity 0.60 (0.46-0.73), positive predictive value 0.71 (0.61-0.82), and negative predictive value of 0.24 (0.17-0.32). The Rome III criteria for abdominal pain are not specific enough to rule out organic causes. Alarm symptoms do not differentiate between organic and functional abdominal pain.

  19. Patterns of arm muscle activation involved in octopus reaching movements.

    Science.gov (United States)

    Gutfreund, Y; Flash, T; Fiorito, G; Hochner, B

    1998-08-01

    The extreme flexibility of the octopus arm allows it to perform many different movements, yet octopuses reach toward a target in a stereotyped manner using a basic invariant motor structure: a bend traveling from the base of the arm toward the tip (Gutfreund et al., 1996a). To study the neuronal control of these movements, arm muscle activation [electromyogram (EMG)] was measured together with the kinematics of reaching movements. The traveling bend is associated with a propagating wave of muscle activation, with maximal muscle activation slightly preceding the traveling bend. Tonic activation was occasionally maintained afterward. Correlation of the EMG signals with the kinematic variables (velocities and accelerations) reveals that a significant part of the kinematic variability can be explained by the level of muscle activation. Furthermore, the EMG level measured during the initial stages of movement predicts the peak velocity attained toward the end of the reaching movement. These results suggest that feed-forward motor commands play an important role in the control of movement velocity and that simple adjustment of the excitation levels at the initial stages of the movement can set the velocity profile of the whole movement. A simple model of octopus arm extension is proposed in which the driving force is set initially and is then decreased in proportion to arm diameter at the bend. The model qualitatively reproduces the typical velocity profiles of octopus reaching movements, suggesting a simple control mechanism for bend propagation in the octopus arm.

  20. Jeopardy alarm system - a building technology like any other technology?; Gefahrenmeldetechnik - eine Gebaeudetechnik wie jede andere?

    Energy Technology Data Exchange (ETDEWEB)

    Hegewald, B. [ZVEI, Fachkreis Ueberfall- und Einbruchmeldeanlagen, Frankfurt am Main (Germany)]|[Siemens AG, Muenchen

    1995-12-31

    In chapter 15 of the anthology about building control the jeopardy alarm system technology is described. The following aspects are discussed: importance of the safety, definition of jeopardy alarm systems, application of safety technique, instructions, guidelines and regulations, advantages of safety systems, integration or combination. (BWI) [Deutsch] Kapitel 15 des Sammelbandes ueber Buidling Control ist der Gefahrenmeldetechnik gewidmet. In diesem Zusammenhang wird auf folgende Themenbereiche eingegangen: Stellenwert der Sicherheit; Definition der Gefahrenmeldeanlagen; Einsatz der Sicherungstechnik; Vorschriften, Richtlinien und Bestimmung; Vorteile von Sicherungssystemen; Integration oder Kombination. (BWI)

  1. Guiding Warfare to Reach Sustainable Peace

    DEFF Research Database (Denmark)

    Vestenskov, David; Drewes, Line

    The conference report Guiding Warfare to Reach Sustainable Peace constitutes the primary outcome of the conference It is based on excerpts from the conference presenters and workshop discussions. Furthermore, the report contains policy recommendations and key findings, with the ambition of develo......The conference report Guiding Warfare to Reach Sustainable Peace constitutes the primary outcome of the conference It is based on excerpts from the conference presenters and workshop discussions. Furthermore, the report contains policy recommendations and key findings, with the ambition...... of developing best practices in the education and implementation of IHL in capacity building of security forces....

  2. Do working environment interventions reach shift workers?

    DEFF Research Database (Denmark)

    Nabe-Nielsen, Kirsten; Jørgensen, Marie Birk; Garde, Anne Helene

    2016-01-01

    PURPOSE: Shift workers are exposed to more physical and psychosocial stressors in the working environment as compared to day workers. Despite the need for targeted prevention, it is likely that workplace interventions less frequently reach shift workers. The aim was therefore to investigate whether...... the reach of workplace interventions varied between shift workers and day workers and whether such differences could be explained by the quality of leadership exhibited at different times of the day. METHODS: We used questionnaire data from 5361 female care workers in the Danish eldercare sector...

  3. Assessment of the Plutonium Finishing Plant Criticality Alarm System U.S. Department of Energy Richland Operations Office

    International Nuclear Information System (INIS)

    NIRIDER, L.T.

    2002-01-01

    At the request of the Assistant Manager for Safety and Engineering, the U.S. Department of Energy Richland Operations Office (RL) Engineering Support Division, performed an oversight review of the Plutonium Finishing Plant (PFP) nuclear Criticality Alarm System (CAS). The review was conducted to satisfy requirements and agreements associated with Defense Nuclear Facility Safety Board (DNFSB) Recommendation 2000-2, ''Vital Safety Systems.'' The PFP is managed by Fluor Hanford, Inc. for RL. The field assessment and staff interviews were conducted August 12 through August 19,2002. This was a limited scope assessment that consisted of a review of the nuclear CAS operations, maintenance, and compliance with National Consensus Standards Requirements. The main purpose of the assessment was to determine the adequacy of the existing alarm system and its associated infrastructure to support the PFP facility mission through the remaining facility lifetime. The Review Plan was modeled upon Criteria and Review Approach Documents (CRAD) developed for DNFSB Recommendation 2000-2 reviews conducted across the Hanford Site. Concerns regarding component degradation and failure, increasing numbers of occurrence reports associated with the alarm system, and reliability issues were addressed. Additionally, RL performed a review of the engineering aspects of the CAS including the functions of design authorities and aspects of systems engineering. However, the focus of the assessment was on operations, maintenance, and reliability of the CAS, associated procurement practices, adequacy of safety and engineering policies and procedures, safety documentation, and fundamental engineering practices including training, qualification, and systems engineering. This assessment revealed that the PFP CAS and its associated infrastructure, administrative procedures, and conduct of operations are generally effective. There are no imminent criticality safety issues associated with the operation of the

  4. Simulation of Columbia River Floods in the Hanford Reach

    Energy Technology Data Exchange (ETDEWEB)

    Waichler, Scott R.; Serkowski, John A.; Perkins, William A.; Richmond, Marshall C.

    2017-01-30

    Columbia River water elevations and flows in the Hanford Reach affect the environment and facilities along the shoreline, including movement of contaminants in groundwater, fish habitat, and infrastructure subject to flooding. This report describes the hydraulic simulation of hypothetical flood flows using the best available topographic and bathymetric data for the Hanford Reach and the Modular Aquatic Simulation System in 1 Dimension (MASS1) hydrodynamic model. The MASS1 model of the Hanford Reach was previously calibrated to field measurements of water surface elevations. The current model setup can be used for other studies of flow, water levels, and temperature in the Reach. The existing MASS1 channel geometry and roughness and other model configuration inputs for the Hanford Reach were used for this study, and previous calibration and validation results for the model are reprinted here for reference. The flood flows for this study were simulated by setting constant flow rates obtained from the U.S. Army Corps of Engineers (USACE) for the Columbia, Snake, and Yakima Rivers, and a constant water level at McNary Dam, and then running the model to steady state. The discharge levels simulated were all low-probability events; for example, a 100-year flood is one that would occur on average every 100 years, or put another way, in any given year there is a 1% chance that a discharge of that level or higher will occur. The simulated floods and their corresponding Columbia River discharges were 100-year (445,000 cfs), 500-year (520,000 cfs), and the USACE-defined Standard Project Flood (960,000 cfs). The resulting water levels from the steady-state floods can be viewed as “worst case” outcomes for the respective discharge levels. The MASS1 output for water surface elevations was converted to the North American Vertical Datum of 1988 and projected across the channel and land surface to enable mapping of the floodplain for each scenario. Floodplain maps show that for

  5. Automated analysis of off-line measured gamma-spectra using UniSampo gamma-ray spectrum analysis software including criterias for alarming systems

    International Nuclear Information System (INIS)

    Nikkinen, M.T.

    2005-01-01

    In many laboratories the number of measured routine gamma-spectra can be significant and the labour work to review all the data is time consuming and expensive task. In many cases the routine sample does not normally contain radiation above a detectable level, and still the review of the spectra has to be performed. By introducing simple rules for emerging conditions, the review work can be significantly reduced. In one case the need to review the environmental measurement spectra was reduced to less than 1% compared to the original need, which in turn made the review personnel available for more useful functions. Using the UniSampo analysis system, the analysis results of spectra that are causing alarming conditions can be transmitted via e-mail to any address. Some systems are even equipped with the capability to forward these results to hand-portable telephones or pagers. This is a very practical solution for automated environmental monitoring, when the sample spectra are collected automatically and transmitted to central computer for further analysis. Set up of an automatic analysis system, rules for the emerging conditions, technical solutions for an automated alarming system and a generic hypothesis test for the alarming system developed for UniSampo analysis software are described. (author)

  6. Constant False Alarm Rate (CFAR) Autotrend Evaluation Report

    Science.gov (United States)

    2011-12-01

    represent a level of uncertainty in the performance analysis. The performance analysis produced the following Key Performance Indicators ( KPIs ) as...Identity KPI Key Performance Indicator MooN M-out-of-N MSPU Modernized Signal Processor Unit NFF No Fault Found PAT Parameter Allocation Table PD

  7. Design of alarm systems in Swedish nuclear power plants; Utformning av larmsystem i svenska kaernkraftverk

    Energy Technology Data Exchange (ETDEWEB)

    Thunberg, Anna; Osvalder, Anna-Lisa (Dept. of Product and Production Development, Chalmers Univ. of Technology, Goeteborg (Sweden))

    2008-04-15

    Research within the area of improving alarm system design and performance has mainly focused on new alarm systems. However, smaller modernisations of legacy systems are more common in the Swedish nuclear industry than design of totally new systems. This imposes problems when the new system should function together with the old system. This project deals with the special concerns raised by modernisation projects. The objective of the project has been to increase the understanding of the relationship between the operator's performance and the design of the alarm system. Of major concern has been to consider the cognitive abilities of the operator, different operator roles and work situations, and varying need of information. The aim of the project has been to complement existing alarm design guidance and to develop user-centred alarm design concepts. Different case studies have been performed in several industry sectors (nuclear, oil refining, pulp and paper, aviation and medical care) to identify best practice. Several empirical studies have been performed within the nuclear area to investigate the operator's need of information, performance and workload in different operating modes. The aspect of teamwork has also been considered. The analyses show that the operator has different roles in different work situations which affect both the type of information needed and how the information is processed. In full power operation, the interaction between the operator and the alarm system is driven by internal factors and the operator tries to maintain high situation awareness by actively searching for information. The operator wants to optimise the process and need detailed information with possibilities to follow-up and get historical data. In disturbance management, the operator is more dependent on external information presented by the alarm system. The new compilation of alarm guidance is based on the operator's varying needs in different working

  8. Reaching Reluctant Students: Insights from Torey Hayden.

    Science.gov (United States)

    Marlowe, Mike

    1999-01-01

    Illustrates principles of reaching students who fight or avoid adults by using examples drawn from the writings of Torey Hayden. Presents ten concepts that can serve as guidelines for building relationships with resistant children, and gives excerpts from Hayden's works to illustrate each concept. Demonstrates how books provide teachers with…

  9. ATLAS Barrel Toroid magnet reached nominal field

    CERN Multimedia

    2006-01-01

     On 9 November the barrel toroid magnet reached its nominal field of 4 teslas, with an electrical current of 21 000 amperes (21 kA) passing through the eight superconducting coils as shown on this graph

  10. Development and Application of a Clinical Microsystem Simulation Methodology for Human Factors-Based Research of Alarm Fatigue.

    Science.gov (United States)

    Kobayashi, Leo; Gosbee, John W; Merck, Derek L

    2017-07-01

    (1) To develop a clinical microsystem simulation methodology for alarm fatigue research with a human factors engineering (HFE) assessment framework and (2) to explore its application to the comparative examination of different approaches to patient monitoring and provider notification. Problems with the design, implementation, and real-world use of patient monitoring systems result in alarm fatigue. A multidisciplinary team is developing an open-source tool kit to promote bedside informatics research and mitigate alarm fatigue. Simulation, HFE, and computer science experts created a novel simulation methodology to study alarm fatigue. Featuring multiple interconnected simulated patient scenarios with scripted timeline, "distractor" patient care tasks, and triggered true and false alarms, the methodology incorporated objective metrics to assess provider and system performance. Developed materials were implemented during institutional review board-approved study sessions that assessed and compared an experimental multiparametric alerting system with a standard monitor telemetry system for subject response, use characteristics, and end-user feedback. A four-patient simulation setup featuring objective metrics for participant task-related performance and response to alarms was developed along with accompanying structured HFE assessment (questionnaire and interview) for monitor systems use testing. Two pilot and four study sessions with individual nurse subjects elicited true alarm and false alarm responses (including diversion from assigned tasks) as well as nonresponses to true alarms. In-simulation observation and subject questionnaires were used to test the experimental system's approach to suppressing false alarms and alerting providers. A novel investigative methodology applied simulation and HFE techniques to replicate and study alarm fatigue in controlled settings for systems assessment and experimental research purposes.

  11. Expert system for the real-time management of alarms in an electric grid. Un systeme expert pour la gestion en temps reel des alarmes dans un reseau electrique

    Energy Technology Data Exchange (ETDEWEB)

    Girouard, P.

    1987-01-01

    A model-base expert system is presented for the processing of alarm messages in power networks' control centers. This alarm processor, called GESTAL, generates concise diagnoses identifying the causes(s) of network disturbances and describing the operation of the protection and alarm systems which operate to isolate faulty components. Essentially, the alarm processor is an agenda driven expert system which is composed of a model of the power network's alarm and protection systems as well as several sets of rules. Based on the nextwork model, graph structures are constructed which represent relationships between alarms as they are received. These graph structures are then analysed in order to diagnose the disturbances(s). The alarm processor has been developed using the ART 3.0 programming language in a Symbolics Lisp-machine environment. Background material on expert systems, electric power networks, and alarm processing software is included, as well as a description of GESTAL'S functional characteristics and design. 51 refs., 28 figs.

  12. The Role of Social Influence on How Residence Hall Inhabitants Respond to Fire Alarms

    Science.gov (United States)

    Leytem, Michael; Stark, Emily

    2016-01-01

    College resident halls pose a threat for a catastrophic event in the case of fire, but little research has examined potential influences on students' responses to fire alarms, particularly the role of social influence in affecting their behaviors. In the current study, residence hall inhabitants reported their knowledge about fire safety, their…

  13. Assessment of predation risk through conspecific alarm odors by spiny lobsters

    Science.gov (United States)

    2009-01-01

    Strong “alarm odors” emanating from lethally injured conspecifics may indicate an imminent risk of predation to spiny lobsters. In laboratory trials,1 strong conspecific alarm odors elicited avoidance in Panulirus argus, a highly gregarious species that displays collective defense behavior, but not in Panulirus guttatus, a species that tends to aggregate when reproductive activity is high (spring) but not when it is low (late summer) and does not display collective defensive behavior. To reduce predation risk, however, lobsters may autotomize limbs, thus sustaining nonlethal injuries. I tested the response of these lobsters to scents emanating from intact, lethally-injured and non-lethally injured conspecifics. In P. argus, these scents elicited, respectively, attraction, avoidance and a random response, suggesting that, in P. argus, avoidance of conspecific alarm odors depends on their strength. In contrast, P. guttatus lobsters responded at random to scents of lethally injured conspecifics and showed a similar response to scents of intact and non-lethally injured conspecifics in the spring (attraction) and in the summer (random), reflecting the more cryptic defensive behavior of this species. Therefore, both species use conspecific alarm odors for risk-assessment, but each responds to these cues in the most effective way to reduce its risk of predation. PMID:19721871

  14. Assessment of predation risk through conspecific alarm odors by spiny lobsters: How much is too much?

    Science.gov (United States)

    Briones-Fourzán, Patricia

    2009-07-01

    Strong "alarm odors" emanating from lethally injured conspecifics may indicate an imminent risk of predation to spiny lobsters. In laboratory trials,1 strong conspecific alarm odors elicited avoidance in Panulirus argus, a highly gregarious species that displays collective defense behavior, but not in Panulirus guttatus, a species that tends to aggregate when reproductive activity is high (spring) but not when it is low (late summer) and does not display collective defensive behavior. To reduce predation risk, however, lobsters may autotomize limbs, thus sustaining nonlethal injuries. I tested the response of these lobsters to scents emanating from intact, lethally-injured and non-lethally injured conspecifics. In P. argus, these scents elicited, respectively, attraction, avoidance and a random response, suggesting that, in P. argus, avoidance of conspecific alarm odors depends on their strength. In contrast, P. guttatus lobsters responded at random to scents of lethally injured conspecifics and showed a similar response to scents of intact and non-lethally injured conspecifics in the spring (attraction) and in the summer (random), reflecting the more cryptic defensive behavior of this species. Therefore, both species use conspecific alarm odors for risk-assessment, but each responds to these cues in the most effective way to reduce its risk of predation.

  15. Automatic diagnosis of alarms: a system to improve operator emergency response

    International Nuclear Information System (INIS)

    Olson, H.P.; Gimmy, K.L.; Nomm, E.; Finley, R.H.

    1980-01-01

    A system is being developed at the Savannah River Plant to help reactor operators respond to multiple alarms in a developing incident situation. The need for such systems has becme evident in recent years, particularly after the Three Mile Island incident

  16. Predictive value of the official cancer alarm symptoms in general practice

    DEFF Research Database (Denmark)

    Krasnik Huggenberger, Ivan; Andersen, John Sahl

    2015-01-01

    Introduction: The objective of this study was to investigate the evidence for positive predictive value (PPV) of alarm symptoms and combinations of symptoms for colorectal cancer, breast cancer, prostate cancer and lung cancer in general practice. Methods: This study is based on a literature search...

  17. Honey Bees Modulate Their Olfactory Learning in the Presence of Hornet Predators and Alarm Component.

    Directory of Open Access Journals (Sweden)

    Zhengwei Wang

    Full Text Available In Southeast Asia the native honey bee species Apis cerana is often attacked by hornets (Vespa velutina, mainly in the period from April to November. During the co-evolution of these two species honey bees have developed several strategies to defend themselves such as learning the odors of hornets and releasing alarm components to inform other mates. However, so far little is known about whether and how honey bees modulate their olfactory learning in the presence of the hornet predator and alarm components of honey bee itself. In the present study, we test for associative olfactory learning of A. cerana in the presence of predator odors, the alarm pheromone component isopentyl acetate (IPA, or a floral odor (hexanal as a control. The results show that bees can detect live hornet odors, that there is almost no association between the innately aversive hornet odor and the appetitive stimulus sucrose, and that IPA is less well associated with an appetitive stimulus when compared with a floral odor. In order to imitate natural conditions, e.g. when bees are foraging on flowers and a predator shows up, or alarm pheromone is released by a captured mate, we tested combinations of the hornet odor and floral odor, or IPA and floral odor. Both of these combinations led to reduced learning scores. This study aims to contribute to a better understanding of the prey-predator system between A. cerana and V. velutina.

  18. CRAW: An expert system for nuclear reactor cover gas alarm analysis

    International Nuclear Information System (INIS)

    Zimmerman, B.D.; Rawlins, J.A.

    1987-08-01

    CRAW is an expert system designed to distinguish fuel pin rupture at the Fast Flux Test Facility. The software is used to evaluate tag gases detected by the alarm system and access operational consequences of a rupture. The ability to distinguish fuel rod as opposed to fuel pin rupture is included. The software is implemented on an IBM-PC

  19. Decree 53/1963 of 10 January on the establishment of a radioactivity alarm network

    International Nuclear Information System (INIS)

    1963-01-01

    This Decree provides for the setting up of a radioactivity alarm network by the Directorate General for Public Protection to prevent hazards arising from possible incidents originating in nuclear installations. The Duty stations are located in specified police stations and with fire brigades within a radius ranging from 15 to 20 km from planned reactors or nuclear power plants. (NEA) [fr

  20. AN APPROACH TO ALLEVIATE THE FALSE ALARM IN BUILDING CHANGE DETECTION FROM URBAN VHR IMAGE

    Directory of Open Access Journals (Sweden)

    J. Chen

    2016-06-01

    Full Text Available Building change detection from very-high-resolution (VHR urban remote sensing image frequently encounter the challenge of serious false alarm caused by different illumination or viewing angles in bi-temporal images. An approach to alleviate the false alarm in urban building change detection is proposed in this paper. Firstly, as shadows casted by urban buildings are of distinct spectral and shape feature, it adopts a supervised object-based classification technique to extract them in this paper. Secondly, on the opposite direction of sunlight illumination, a straight line is drawn along the principal orientation of building in every extracted shadow region. Starting from the straight line and moving toward the sunlight direction, a rectangular area is constructed to cover partial shadow and rooftop of each building. Thirdly, an algebra and geometry invariant based method is used to abstract the spatial topological relationship of the potential unchanged buildings from all central points of the rectangular area. Finally, based on an oriented texture curvature descriptor, an index is established to determine the actual false alarm in building change detection result. The experiment results validate that the proposed method can be used as an effective framework to alleviate the false alarm in building change detection from urban VHR image.

  1. Evaluation of coverage of enriched UF6 cylinder storage lots by existing criticality accident alarms

    International Nuclear Information System (INIS)

    Lee, B.L. Jr.; Dobelbower, M.C.; Woollard, J.E.; Sutherland, P.J.; Tayloe, R.W. Jr.

    1995-03-01

    The Portsmouth Gaseous Diffusion Plant (PORTS) is leased from the US Department of Energy (DOE) by the United States Enrichment Corporation (USEC), a government corporation formed in 1993. PORTS is in transition from regulation by DOE to regulation by the Nuclear Regulatory Commission (NRC). One regulation is 10 CFR Part 76.89, which requires that criticality alarm systems be provided for the site. PORTS originally installed criticality accident alarm systems in all building for which nuclear criticality accidents were credible. Currently, however, alarm systems are not installed in the enriched uranium hexafluoride (UF 6 ) cylinder storage lots. This report analyzes and documents the extent to which enriched UF 6 cylinder storage lots at PORTS are covered by criticality detectors and alarms currently installed in adjacent buildings. Monte Carlo calculations are performed on simplified models of the cylinder storage lots and adjacent buildings. The storage lots modelled are X-745B, X-745C, X745D, X-745E, and X-745F. The criticality detectors modelled are located in building X-343, the building X-344A/X-342A complex, and portions of building X-330 (see Figures 1 and 2). These criticality detectors are those located closest to the cylinder storage lots. Results of this analysis indicate that the existing criticality detectors currently installed at PORTS are largely ineffective in detecting neutron radiation from criticality accidents in most of the cylinder storage lots at PORTS, except sometimes along portions of their peripheries

  2. Alarm signs and antibiotic prescription in febrile children in primary care : an observational cohort study

    NARCIS (Netherlands)

    Elshout, Gijs; van Ierland, Yvette; Bohnen, Arthur M.; de Wilde, Marcel; Oostenbrink, Rianne; Moll, Henriette A.; Berger, Marjolein Y.

    Background Although fever in children is often self-limiting, antibiotics are frequently prescribed for febrile illnesses. GPs may consider treating serious infections by prescribing antibiotics. Aim To examine whether alarm signs and/or symptoms for serious infections are related to antibiotic

  3. Validation of the Rome III criteria and alarm symptoms for recurrent abdominal pain in children

    NARCIS (Netherlands)

    Gijsbers, Carolien F. M.; Benninga, Marc A.; Schweizer, Joachim J.; Kneepkens, C. M. Frank; Vergouwe, Yvonne; Büller, Hans A.

    2014-01-01

    Rome criteria were formulated to define functional gastrointestinal disorders (Rome III criteria, 2006) excluding organic diagnoses when alarm symptoms were absent. The aims of the study were to validate the Rome III criteria as to their capacity to differentiate between organic and functional

  4. Specific and unspecific gynecological alarm symptoms -prevalence estimates in different age groups

    DEFF Research Database (Denmark)

    Balasubramaniam, Kirubakaran; Ravn, Pernille; Larsen, Pia V

    2015-01-01

    OBJECTIVE: To determine prevalence estimates of gynecological alarm symptoms in different age groups and to describe common patterns of gynecological symptoms. DESIGN: Web-based cross-sectional survey study. SETTING: Nationwide in Denmark. POPULATION: A random sample of 51 090 women aged 20 years...

  5. From nestling calls to fledgling silence: adaptive timing of change in response to aerial alarm calls.

    Science.gov (United States)

    Magrath, Robert D; Platzen, Dirk; Kondo, Junko

    2006-09-22

    Young birds and mammals are extremely vulnerable to predators and so should benefit from responding to parental alarm calls warning of danger. However, young often respond differently from adults. This difference may reflect: (i) an imperfect stage in the gradual development of adult behaviour or (ii) an adaptation to different vulnerability. Altricial birds provide an excellent model to test for adaptive changes with age in response to alarm calls, because fledglings are vulnerable to a different range of predators than nestlings. For example, a flying hawk is irrelevant to a nestling in a enclosed nest, but is dangerous to that individual once it has left the nest, so we predict that young develop a response to aerial alarm calls to coincide with fledging. Supporting our prediction, recently fledged white-browed scrubwrens, Sericornis frontalis, fell silent immediately after playback of their parents' aerial alarm call, whereas nestlings continued to calling despite hearing the playback. Young scrubwrens are therefore exquisitely adapted to the changing risks faced during development.

  6.   Combination treatment of monosymptomatic enuresis nocturna with alarm and desmopressin

    DEFF Research Database (Denmark)

    Kamperis, Konstantinos; Hagstrøm, Søren; Rittig, Søren

    2006-01-01

    children had completed the diagnostic procedures of our center comprising 2-week home recordings, desmopressin titration, uroflowmetry and urinalysis. The latest ICCS standardization was used for characterizations. All children were treated with the enuresis alarm alone or in combination with desmopressin...

  7. Manifold adaptation for constant false alarm rate ship detection in South African oceans

    CSIR Research Space (South Africa)

    Schwegmann, CP

    2015-04-01

    Full Text Available prescreening method. It uses a scalar threshold value to determine how bright a pixel needs to be in order to be classified as a ship and thus inversely how many false alarms are permitted. This paper presents by a method of converting the scalar threshold...

  8. Alarm radiation dosimeter with improved integrating pulse ionization chamber and high voltage supply

    International Nuclear Information System (INIS)

    Borkowski, C.J.; Rochelle, J.M.

    1975-01-01

    An alarm dosimeter is described which features an improved integrating pulse ionization chamber of the type containing an hermetically sealed gas diode. Improved operation and miniaturization of the chamber are made possible by a ringing choke converter high voltage supply having a ripple-type output that insures discharge of the gas diode. (author)

  9. Alarm signs and antibiotic prescription in febrile children in primary care: An observational cohort study

    NARCIS (Netherlands)

    G. Elshout (Gijs); Y. van Ierland (Yvette); A.M. Bohnen (Arthur); M. de Wilde (Marcel); R. Oostenbrink (Rianne); H.A. Moll (Henriëtte); M.Y. Berger (Marjolein)

    2013-01-01

    textabstractBackground Although fever in children is often self-limiting, antibiotics are frequently prescribed for febrile illnesses. GPs may consider treating serious infections by prescribing antibiotics. Aim To examine whether alarm signs and/or symptoms for serious infections are related to

  10. Predator guild does not influence orangutan alarm call rates and combinations

    NARCIS (Netherlands)

    Lameira, A.R.; de Vries, H.; Hardus, M.E.; Hall, C.P.A.; Mitra-Setia, T.; Spruijt, B.M.; Kershenbaum, A.; Sterck, E.H.M.; van Noordwijk, M.; van Schaick, C.; Wich, S.A.

    2013-01-01

    Monkey alarm calls have shown that in the primate clade, combinatorial rules in acoustic communication are not exclusive to humans. A recent hypothesis suggests that the number of different call combinations in monkeys increases with increased number of predator species. However, the existence of

  11. Reevaluation of the PMS alarm set-points in OPR1000

    International Nuclear Information System (INIS)

    Roh, Kyung Ho; Yang, Sung Tae; Jung, Sung In

    2011-01-01

    In Optimized Power Reactor 1000 (OPR1000), the common alarm of the plant monitoring system (PMS), which is related to the channel-to-channel core protection calculator (CPC), experiences frequent deviations in the departure from nucleate boiling ratio (DNBR) and the local power density (LPD) between the middle of the cycle and the end of the cycle. Because the channel-to-channel CPC causes deviations in the values of the DNBR and LPD, the increase in the CPC input variables exceeds the alarm set-points. The CPC DNBR and LPD are defined as follows: Deviation = the average of four CPC channels of the LPD and DNBR minus the value of each CPC channel LPD and DNBR. In this paper, we report on a review by the Korea Hydro and Nuclear Power Co., Ltd. (KHNP) regarding the suitability of the alarm set-points for the channel-tochannel deviations of the CPC DNBR and LPD. The set-points were revaluated in light of operational experience and the case of Palo Verde (which is the reference model of OPR1000). The KHNP consequently revised the relevant procedures, as well as and the PMS alarm set-points, as part of its follow-up action

  12. Alarm communication: a new function for the scent-gland secretion in harvestmen (Arachnida: Opiliones)

    Science.gov (United States)

    Machado, Glauco; Bonato, Vinícius; Oliveira, Paulo

    2002-05-01

    Most harvestmen are nocturnal, nonacoustical, and nonvisual arthropods. They have a pair of exocrine glands on the cephalothorax that produce defensive volatile secretions. We investigated in the field the possible alarm effect of these secretions in the gregarious harvestman Goniosoma aff. proximum. A cotton swab soaked with the species' own exudate (treatment), or with water (control), was held 1-2 cm from the center of harvestmen aggregations. The results showed that the gland secretion elicits an alarm response in Goniosoma: whereas 73.3% of the aggregations dispersed after being stimulated with the gland exudate, only 3.3% responded to the water control. Respondent groups are larger than non-respondent groups, and the time of reaction to the secretion was inversely related to group size. This is the first demonstration of a chemically-mediated alarm effect in harvestmen. The alarm response in gregarious harvestmen has possibly evolved as a by-product of a primarily defensive reaction in the context of predator avoidance. The discovery of this novel function of scent-gland secretion is meaningful in view of the widespread occurrence of gregarious habit among species of the order Opiliones.

  13. CSER 95-003: Exemption from Criticality Alarm System requirement for 232-Z building

    International Nuclear Information System (INIS)

    Nirider, L.T.; Miller, E.M.

    1995-01-01

    This CSER establishes an exemption for 232-Z from the requirement for a Criticality Alarm System, because the formation of a critical configuration is not a credible event for any circumstance involving the cleaning out and removal of the Burning Hood and associated equipment

  14. Inactivation of Parietal Reach Region Affects Reaching But Not Saccade Choices in Internally Guided Decisions.

    Science.gov (United States)

    Christopoulos, Vassilios N; Bonaiuto, James; Kagan, Igor; Andersen, Richard A

    2015-08-19

    The posterior parietal cortex (PPC) has traditionally been considered important for awareness, spatial perception, and attention. However, recent findings provide evidence that the PPC also encodes information important for making decisions. These findings have initiated a running argument of whether the PPC is critically involved in decision making. To examine this issue, we reversibly inactivated the parietal reach region (PRR), the area of the PPC that is specialized for reaching movements, while two monkeys performed a memory-guided reaching or saccade task. The task included choices between two equally rewarded targets presented simultaneously in opposite visual fields. Free-choice trials were interleaved with instructed trials, in which a single cue presented in the peripheral visual field defined the reach and saccade target unequivocally. We found that PRR inactivation led to a strong reduction of contralesional choices, but only for reaches. On the other hand, saccade choices were not affected by PRR inactivation. Importantly, reaching and saccade movements to single instructed targets remained largely intact. These results cannot be explained as an effector-nonspecific deficit in spatial attention or awareness, since the temporary "lesion" had an impact only on reach choices. Hence, the PPR is a part of a network for reach decisions and not just reach planning. There has been an ongoing debate on whether the posterior parietal cortex (PPC) represents only spatial awareness, perception, and attention or whether it is also involved in decision making for actions. In this study we explore whether the parietal reach region (PRR), the region of the PPC that is specialized for reaches, is involved in the decision process. We inactivated the PRR while two monkeys performed reach and saccade choices between two targets presented simultaneously in both hemifields. We found that inactivation affected only the reach choices, while leaving saccade choices intact

  15. Sequence and batch language programs and alarm-related ``C`` programs for the 242-A MCS. Revision 2

    Energy Technology Data Exchange (ETDEWEB)

    Berger, J.F.

    1995-03-01

    A Distributive Process Control system was purchased by Project B-534, ``242-A Evaporator/Crystallizer Upgrades``. This control system, called the Monitor and Control System (MCS), was installed in the 242-A Evaporator located in the 200 East Area. The purpose of the MCS is to monitor and control the Evaporator and monitor a number of alarms and other signals from various Tank Farm facilities. Applications software for the MCS was developed by the Waste Treatment Systems Engineering (WTSE) group of Westinghouse. The standard displays and alarm scheme provide for control and monitoring, but do not directly indicate the signal location or depict the overall process. To do this, WTSE developed a second alarm scheme which uses special programs, annunciator keys, and process graphics. The special programs are written in two languages; Sequence and Batch Language (SABL), and ``C`` language. The WTSE-developed alarm scheme works as described below: SABL relates signals and alarms to the annunciator keys, called SKID keys. When an alarm occurs, a SABL program causes a SKID key to flash, and if the alarm is of yellow or white priority then a ``C`` program turns on an audible horn (the D/3 system uses a different audible horn for the red priority alarms). The horn and flashing key draws the attention of the operator.

  16. Persons with Mild or Moderate Alzheimer's Disease Learn to Use Urine Alarms and Prompts to Avoid Large Urinary Accidents

    Science.gov (United States)

    Lancioni, Giulio E.; Singh, Nirbhay N.; O'Reilly, Mark F.; Sigafoos, Jeff; Bosco, Andrea; Zonno, Nadia; Badagliacca, Francesco

    2011-01-01

    This study assessed whether three patients with Alzheimer's disease could learn to use urine alarms and caregivers' prompts to eliminate large urinary accidents. As soon as the patient began to release urine, the alarm system presented auditory and vibratory signals. In relation to those signals, the caregiver would prompt/encourage the patient to…

  17. Alarming signs and symptoms in febrile children in primary care: an observational cohort study in The Netherlands.

    Directory of Open Access Journals (Sweden)

    Gijs Elshout

    Full Text Available CONTEXT: Febrile children in primary care have a low risk for serious infection. Although several alarming signs and symptoms are proposed to have predictive value for serious infections, most are based on research in secondary care. The frequency of alarming signs/symptoms has not been established in primary care; however, in this setting differences in occurrence may influence their predictive value for serious infections. OBJECTIVE: To determine the frequency of alarming signs/symptoms in febrile children in primary care. DESIGN: Observational cohort study. Clinical information was registered in a semi-structured way and manually recoded. SETTING: General practitioners' out-of-hours service. SUBJECTS: Face-to-face patient contacts concerning children (aged ≤16 years with fever were eligible for inclusion. MAIN OUTCOME MEASURES: Frequency of 18 alarming signs and symptoms as reported in the literature. RESULTS: A total of 10,476 patient contacts were included. The frequency of alarming signs/symptoms ranged from n = 1 (ABC instability; 40°C as reported by the parents; 12.9% to 8,647 contacts (parental concern; 82.5%. CONCLUSION: Although the prevalence of specific alarming signs/symptoms is low in primary care, ≥50% of children have one or more alarming signs/symptoms. There is a need to determine the predictive value of alarming signs/symptoms not only for serious infections in primary care, but as well for increased risk of a complicated course of the illness.

  18. The nature of alarm communication in Constrictotermes cyphergaster (Blattodea: Termitoidea: Termitidae): the integration of chemical and vibroacoustic signals

    Czech Academy of Sciences Publication Activity Database

    Cristaldo, P. F.; Jandák, V.; Kutalová, Kateřina; Rodrigues, V. B.; Brothánek, M.; Jiříček, O.; DeSouza, O.; Šobotník, J.

    2015-01-01

    Roč. 4, č. 12 (2015), s. 1649-1659 ISSN 2046-6390 Institutional support: RVO:61388963 Keywords : alarm communication * alarm pheromone * defence * Isoptera * Nasutitermitinae * vibroacoustic communication Subject RIV: EG - Zoology Impact factor: 2.135, year: 2015 http://bio.biologists.org/content/biolopen/4/12/1649.full.pdf

  19. Does workplace health promotion reach shift workers?

    DEFF Research Database (Denmark)

    Nabe-Nielsen, Kirsten; Garde, Anne Helene; Clausen, Thomas

    2015-01-01

    OBJECTIVES: One reason for health disparities between shift and day workers may be that workplace health promotion does not reach shift workers to the same extent as it reaches day workers. This study aimed to investigate the association between shift work and the availability of and participation...... in workplace health promotion. METHODS: We used cross-sectional questionnaire data from a large representative sample of all employed people in Denmark. We obtained information on the availability of and participation in six types of workplace health promotion. We also obtained information on working hours, ie......). RESULTS: In the general working population, fixed evening and fixed night workers, and employees working variable shifts including night work reported a higher availability of health promotion, while employees working variable shifts without night work reported a lower availability of health promotion...

  20. Performance reach in the LHC for 2012

    International Nuclear Information System (INIS)

    Arduini, G.

    2012-01-01

    Based on the 2011 experience and Machine Development study results, the performance reach of the LHC with 25 and 50 ns beams will be addressed for operation at 3.5 and 4 TeV. The possible scrubbing scenarios and potential intensity limitations resulting from vacuum, heating will be taken into account wherever possible. The paper mainly covers the performance of the two high luminosity regions in IR1 and IR5. (author)

  1. Comparison of electronic digital alarm dosimeter with TLD

    International Nuclear Information System (INIS)

    Kumar, Pankaj; Pandey, J.P.N.; Shinde, A M.; Purohit, R.G.; Sarkar, P.K.

    2012-01-01

    Control of exposure of radiation workers on day to day basis has been made easy by use of semiconductor based electronic digital dosimeter. Additional dose constraints of 10 mSv for occupational radiation workers have made it essential to use such type of digital personal monitoring devices. In addition to conventional ionisation chamber based direct reading dosimeters, additional 35 semiconductor based digital dosimeters model MGP DMC 2000 S were used for the monitoring of personal exposure of radiation workers in a spent fuel reprocessing plant. Though better least count and good performance over a wide range of dose rate are claimed by the manufacture, before making use of such dosimeter on large scale, validation of its performance is required to be checked. In this paper, an effort is made to determine the performance of digital dosimeters, by exposing these digital dosimeters in combination with TLDs at different radiation levels and obtained results were compared and analysed

  2. Associations Between the Self-Reported Frequency of Hearing Chemical Alarms in Theater and Visuospatial Function in Gulf War Veterans.

    Science.gov (United States)

    Chao, Linda L

    2016-10-01

    The aim of this study was to examine the relationship between the self-reported frequencies of hearing chemical alarms during deployment and visuospatial function in Gulf War (GW) veterans. The relationship between the self-reported frequency of hearing chemical alarms, neurobehavioral, and volumetric brain imaging data was examined with correlational, regression, and mediation analyses. The self-reported frequency of hearing chemical alarms was inversely associated with and significantly predicted performance on a visuospatial task (ie, Block Design) over and above potentially confounding variables, including concurrent, correlated GW-related exposures. This effect was partially mediated by the relationship between hearing chemical alarms and lateral occipital cortex volume. Exposure to substances that triggered chemical alarms during GW deployment likely had adverse effects on veterans' brain structure and function, warranting further investigation of whether these GW veterans are at an increased risk for dementia.

  3. Development of electric power transmission line anomaly approaching alarm system. Sodensen ijo sekkin keiho sochi no kaihatsu

    Energy Technology Data Exchange (ETDEWEB)

    Kusaba, Masaro; Tabata, Takatoshi; Miyazaki, Shusuke [Kyushu Electric Power Co., Inc., Fukuoka (Japan)

    1989-03-30

    In order to plan preventive measures from the occurrence of electric power transmission line accident, a power transmission line anomaly approaching alarm system was developed to automatically alarm the operators and surveillants. The surveillance system, composed of camera, surveillance unit proper and alarm receiving portion, generates alarm and indicate hazard to the operators, when animal invades a hazardous area designated on the monitor. Wireless telegraph for the alarm transmission use used frequency around a 322MHz band with the Radio Ray law and preventive measures from noise, taken into consideration. The prototype is characterized by unnecessary ancillary fitting to the surveillance object, rare erroneous operations and high reliability. While the hazardous area can be simply designated, changed and released, as made by volume on the control panel. It is moreover easy to designate, as done by confirming the surveillance object on the display. It is also possible in multidimensional surveillance and easy to use on site. 7 figs.

  4. Factors influencing when intensive care unit nurses go to the bedside to investigate patient related alarms: A descriptive qualitative study.

    Science.gov (United States)

    Despins, Laurel A

    2017-12-01

    This study examines what prompts the intensive care unit (ICU) nurse to go to the patient's bedside to investigate an alarm and the influences on the nurse's determination regarding how quickly this needs to occur. A qualitative descriptive design guided data collection and analysis. Individual semi-structured interviews were conducted. Thematic analysis guided by the Patient Risk Detection Theoretical Framework was applied to the data. Four specialty intensive care units in an academic medical center. ICU nurses go the patient's bedside in response to an alarm to catch patient deterioration and avert harm. Their determination of the immediacy of patient risk and their desire to prioritize their bedside investigations to true alarms influences how quickly they proceed to the bedside. Ready visual access to physiological data and waveform configurations, experience, teamwork, and false alarms are important determinants in the timing of ICU nurses' bedside alarm investigations. Copyright © 2017 Elsevier Ltd. All rights reserved.

  5. Replacement of the level control of draining tanks MSRS and powered water heaters with the OVATION system in Asco NPP

    International Nuclear Information System (INIS)

    Serrano Jimenez, J.

    2012-01-01

    The current MSR drains and heaters tanks level control is local control individual, pneumatic and without action from Control room. The system has level switches for the generation of alarms, isolations and shots of bombs. Single control room operators have level alarms, final race of valves of control and indication of temperature and pressure of some tanks.

  6. Alarming of exchange rate crisis: A risk management approach

    Directory of Open Access Journals (Sweden)

    Da Zhao

    2016-05-01

    Full Text Available Recently, with increasing volatility of foreign exchange rate, risk management becomes more and more important not only for multinational companies and individuals but also for central governments. This paper attempts to build an econometrics model so as to forecast and manage risks in foreign exchange market, especially during the eve of turbulent periods. By following McNeil and Frey’s (2000 two stage approach called conditional EVT to estimate dynamic VaR commonly used in stock and insurance markets, we extend it by applying a more general asymmetric ARMA-GARCH model to analyze daily foreign exchange dollar-denominated trading data from four countries of different development levels across Asia and Europe for a period of more than 10 years from January 03, 2005 to May 29, 2015, which is certainly representative of global markets. Conventionally, different kinds of backtesting methods are implemented ultimately to evaluate how well the model behaves. Inspiringly, test results show that by taking several specific characteristics (including fat-tails, asymmetry and long-range dependence of the foreign exchange market return data into consideration, the violation ratio of out-of-sample data can be forecasted very well for both fixed and flexible foreign exchange regimes. Moreover, all of the violations are evenly distributed along the whole period which indicates another favorable property of our model. Meanwhile, we find evidence of asymmetry volatility in all of the studied foreign exchange markets even though the magnitudes of the most of them are weak

  7. Optimized Vibration Chamber for Landslide Sensory and Alarm System

    Science.gov (United States)

    Ismail, Eliza Sabira Binti; Hadi Habaebi, Mohamed; Daoud, Jamal I.; Rafiqul Islam, Md

    2017-11-01

    Landslide is one of natural hazard that is not unfamiliar disaster in Malaysia. Malaysia has experienced this disaster many times since 1969. This natural hazard has become a major research concern for Malaysian government when many people were injured badly and even had been killed. Many previous research works published in the open literature aimed at designing a system that could detect landslide in early stage before the landslide becomes catastrophic. This paper presents the early works on a major work-in-progress landslide early warning system for Malaysian environment. The aim of this system is to develop the most efficiently reliable cost-effective system in which slight earth movements are monitored continuously. The challenge this work aims at is to work with a low budget system that produces efficient performance. Hence, the material used is off-the-shelf. Early design optimization results of the vibration sensor used is quite promising detecting the slightest faint tremors, which are amplified using the best vibration chamber available. It is shown that the choice of proper pipe length and diameter dimensions in combination to a gravel to exaggerate the produced higher sensitivity level noise of 5 dB.

  8. Riparian Vegetation Mapping Along the Hanford Reach

    International Nuclear Information System (INIS)

    FOGWELL, T.W.

    2003-01-01

    During the biological survey and inventory of the Hanford Site conducted in the mid-1990s (1995 and 1996), preliminary surveys of the riparian vegetation were conducted along the Hanford Reach. These preliminary data were reported to The Nature Conservancy (TNC), but were not included in any TNC reports to DOE or stakeholders. During the latter part of FY2001, PNNL contracted with SEE Botanical, the parties that performed the original surveys in the mid 1990s, to complete the data summaries and mapping associated with the earlier survey data. Those data sets were delivered to PNNL and the riparian mapping by vegetation type for the Hanford Reach is being digitized during the first quarter of FY2002. These mapping efforts provide the information necessary to create subsequent spatial data layers to describe the riparian zone according to plant functional types (trees, shrubs, grasses, sedges, forbs). Quantification of the riparian zone by vegetation types is important to a number of DOE'S priority issues including modeling contaminant transport and uptake in the near-riverine environment and the determination of ecological risk. This work included the identification of vegetative zones along the Reach by changes in dominant plant species covering the shoreline from just to the north of the 300 Area to China Bar near Vernita. Dominant and indicator species included Agropyron dasytachyudA. smithii, Apocynum cannabinum, Aristida longiseta, Artemisia campestris ssp. borealis var scouleriana, Artemisa dracunculus, Artemisia lindleyana, Artemisia tridentata, Bromus tectorum, Chrysothamnus nauseosus, Coreopsis atkinsoniana. Eleocharis palustris, Elymus cinereus, Equisetum hyemale, Eriogonum compositum, Juniperus trichocarpa, Phalaris arundinacea, Poa compressa. Salk exigua, Scirpus acutus, Solidago occidentalis, Sporobolus asper,and Sporobolus cryptandrus. This letter report documents the data received, the processing by PNNL staff, and additional data gathered in FY2002

  9. System for indicating the level of material in a container

    International Nuclear Information System (INIS)

    Erb, T.L.

    1980-01-01

    In a radiation detecting system for controlling the level of material in a container, the first counter accumulates pulses generated by a geiger tube at a rate related to the level of material and a second counter accumulates clock pulses. A race condition is established between a NAND circuit indicating that the first counter has reached a predetermined total, and a NAND circuit indicating that the second counter has reached a second predetermined total representing a fixed counting interval. The first NAND circuit to respond to its predetermined total actuates a circuit to reset both counters and, if indicative of the material level being below a predetermined minimum, actuates an alarm or operates a control circuit to add material to the container. In the example shown, an additional NAND circuit responds to a different count in the first counter which count in the same time interval corresponds to a higher level, and when material is being added to the container, the race condition is between two NAND circuits. The effect of this is to provide a hysteresis effect preventing the circuit from 'hunting' around one level of material. (author)

  10. Long-reach manipulators for decommissioning

    International Nuclear Information System (INIS)

    Webster, D.A.; Challinor, S.F.

    1993-01-01

    A survey of redundant facilities at Sellafield has identified that in many cases the conventional means of deploying remote handling equipment are not appropriate and that novel means must be employed. However, decommissioning is not a value adding activity and so expensive one off designs must be avoided. The paper will describe BNFL's approach to the synthesis from proprietary parts of a manipulator which can lift 3 te at a horizontal reach of over 5 metres and yet can still perform the dextrous manipulation necessary to remove small items. It will also cover the development of the manipulator control systems and the adaption of commercial handtools to be manipulator friendly. (author)

  11. Luminosity performance reach after LS1

    International Nuclear Information System (INIS)

    Herr, W.

    2012-01-01

    Based on past experience (2010/2011), in particular expected limitations from beam-beam effects, and taking into account the expected beam quality from the LHC injectors, the peak and integrated luminosity at top energy is discussed for different scenarios (e.g. bunch spacing, beta*). In particular it will be shown which are the key parameters to reach the nominal luminosity and it is also shown that peak luminosities two times larger than nominal (or higher) are possible. Possible test in 2012 are discussed

  12. Multiple-Parameter, Low-False-Alarm Fire-Detection Systems

    Science.gov (United States)

    Hunter, Gary W.; Greensburg, Paul; McKnight, Robert; Xu, Jennifer C.; Liu, C. C.; Dutta, Prabir; Makel, Darby; Blake, D.; Sue-Antillio, Jill

    2007-01-01

    Fire-detection systems incorporating multiple sensors that measure multiple parameters are being developed for use in storage depots, cargo bays of ships and aircraft, and other locations not amenable to frequent, direct visual inspection. These systems are intended to improve upon conventional smoke detectors, now used in such locations, that reliably detect fires but also frequently generate false alarms: for example, conventional smoke detectors based on the blockage of light by smoke particles are also affected by dust particles and water droplets and, thus, are often susceptible to false alarms. In contrast, by utilizing multiple parameters associated with fires, i.e. not only obscuration by smoke particles but also concentrations of multiple chemical species that are commonly generated in combustion, false alarms can be significantly decreased while still detecting fires as reliably as older smoke-detector systems do. The present development includes fabrication of sensors that have, variously, micrometer- or nanometer-sized features so that such multiple sensors can be integrated into arrays that have sizes, weights, and power demands smaller than those of older macroscopic sensors. The sensors include resistors, electrochemical cells, and Schottky diodes that exhibit different sensitivities to the various airborne chemicals of interest. In a system of this type, the sensor readings are digitized and processed by advanced signal-processing hardware and software to extract such chemical indications of fires as abnormally high concentrations of CO and CO2, possibly in combination with H2 and/or hydrocarbons. The system also includes a microelectromechanical systems (MEMS)-based particle detector and classifier device to increase the reliability of measurements of chemical species and particulates. In parallel research, software for modeling the evolution of a fire within an aircraft cargo bay has been developed. The model implemented in the software can

  13. Approaching Behaviour Monitor and Vibration Indication in Developing a General Moving Object Alarm System (GMOAS

    Directory of Open Access Journals (Sweden)

    Haiwei Dong

    2013-07-01

    Full Text Available People who suffer from hearing impairment caused by illness, age or extremely noisy environments are constantly in danger of being hit or knocked down by fast moving objects behind them when they have no companion or augmented sensory system to warn them. In this paper, we propose the General Moving Object Alarm System (GMOAS, a system focused on aiding the safe mobility of people under these circumstances. The GMOAS is a wearable haptic device that consists of two main subsystems: (i a moving object monitoring subsystem that uses laser range data to detect and track approaching objects, and (ii an alarm subsystem that warns the user of possibly dangerous approaching objects by triggering tactile vibrations on an “alarm necklace”. For moving object monitoring, we propose a simple yet efficient solution to monitor the approaching behavior of objects. Compared with previous work in motion detection and tracking, we are not interested in specific objects but any type of approaching object that might harm the user. To this extent, we define a boundary in the laser range data where the objects are monitored. Within this boundary a fan-shape grid is constructed to obtain an evenly distributed spatial partitioning of the data. These partitions are efficiently clustered into continuous objects which are then tracked through time using an object association algorithm based on updating a deviation matrix that represents angle, distance and size variations of the objects. The speed of the tracked objects is monitored throughout the algorithm. When the speed of an approaching object surpasses the safety threshold, the alarm necklace is triggered indicating the approaching direction of the fast moving object. The alarm necklace is equipped with three motors that can indicate five directions with respect to the user: left, back, right, left-back and right-back. We performed three types of outdoor experiments (object passing, approaching and crossing that

  14. Correlation of Salivary Glucose Level with Blood Glucose Level in Diabetes Mellitus

    Directory of Open Access Journals (Sweden)

    Arati S. Panchbhai

    2012-07-01

    Full Text Available Objectives: There is alarming rise in number of people with diabetes mellitus over these years. If glucose in saliva is linked to glucose in blood it can be used to detect diabetes mellitus at an early stage. The present study is undertaken with the aim to assess the correlation of salivary glucose level with blood glucose level in people with diabetes mellitus. Material and Methods: For investigations, 2 sets of samples of people with diabetes and the age and sex matched non-diabetic subjects were recruited. The salivary glucose was analyzed in unstimulated whole saliva samples using glucose oxidase method. Pearson’s correlation coefficient test was applied to assess the correlation between salivary glucose level and blood glucose level. Results: The significant (P < 0.05 positive correlation of salivary glucose level and fasting blood glucose level was observed in people with uncontrolled diabetes in both the sets of samples.Conclusions: Although study suggests some potential for saliva as a marker in monitoring of diabetes mellitus, there are many aspects that need clarification before we reach to a conclusion.

  15. Can donated media placements reach intended audiences?

    Science.gov (United States)

    Cooper, Crystale Purvis; Gelb, Cynthia A; Chu, Jennifer; Polonec, Lindsey

    2013-09-01

    Donated media placements for public service announcements (PSAs) can be difficult to secure, and may not always reach intended audiences. Strategies used by the Centers for Disease Control and Prevention's (CDC) Screen for Life: National Colorectal Cancer Action Campaign (SFL) to obtain donated media placements include producing a diverse mix of high-quality PSAs, co-branding with state and tribal health agencies, securing celebrity involvement, monitoring media trends to identify new distribution opportunities, and strategically timing the release of PSAs. To investigate open-ended recall of PSAs promoting colorectal cancer screening, CDC conducted 12 focus groups in three U.S. cities with men and women either nearing age 50 years, when screening is recommended to begin, or aged 50-75 years who were not in compliance with screening guidelines. In most focus groups, multiple participants recalled exposure to PSAs promoting colorectal cancer screening, and most of these individuals reported having seen SFL PSAs on television, in transit stations, or on the sides of public buses. Some participants reported exposure to SFL PSAs without prompting from the moderator, as they explained how they learned about the disease. Several participants reported learning key campaign messages from PSAs, including that colorectal cancer screening should begin at age 50 years and screening can find polyps so they can be removed before becoming cancerous. Donated media placements can reach and educate mass audiences, including millions of U.S. adults who have not been screened appropriately for colorectal cancer.

  16. Efficacy of REACH Forgiveness across cultures.

    Science.gov (United States)

    Lin, Yin; Worthington, Everett L; Griffin, Brandon J; Greer, Chelsea L; Opare-Henaku, Annabella; Lavelock, Caroline R; Hook, Joshua N; Ho, Man Yee; Muller, Holly

    2014-09-01

    This study investigates the efficacy of the 6-hour REACH Forgiveness intervention among culturally diverse undergraduates. Female undergraduates (N = 102) and foreign extraction (46.2%) and domestic (43.8%) students in the United States were randomly assigned to immediate treatment or waitlist conditions. Treatment efficacy and the effect of culture on treatment response were assessed using measures of emotional and decisional forgiveness across 3 time periods. Students in the treatment condition reported greater improvement in emotional forgiveness, but not decisional forgiveness, relative to those in the waitlist condition. Gains were maintained at a 1-week follow-up. Although culture did not moderate the effect of treatment, a main effect of culture on emotional forgiveness and marginally significant interaction effect of culture on decisional forgiveness were found. The REACH Forgiveness intervention was efficacious for college students from different cultural backgrounds when conducted in the United States. However, some evidence may warrant development of culturally adapted forgiveness interventions. © 2014 Wiley Periodicals, Inc.

  17. Bimanual reach to grasp movements after cervical spinal cord injury.

    Directory of Open Access Journals (Sweden)

    Laura Britten

    Full Text Available Injury to the cervical spinal cord results in bilateral deficits in arm/hand function reducing functional independence and quality of life. To date little research has been undertaken to investigate control strategies of arm/hand movements following cervical spinal cord injury (cSCI. This study aimed to investigate unimanual and bimanual coordination in patients with acute cSCI using 3D kinematic analysis as they performed naturalistic reach to grasp actions with one hand, or with both hands together (symmetrical task, and compare this to the movement patterns of uninjured younger and older adults. Eighteen adults with a cSCI (mean 61.61 years with lesions at C4-C8, with an American Spinal Injury Association (ASIA grade B to D and 16 uninjured younger adults (mean 23.68 years and sixteen uninjured older adults (mean 70.92 years were recruited. Participants with a cSCI produced reach-to-grasp actions which took longer, were slower, and had longer deceleration phases than uninjured participants. These differences were exacerbated during bimanual reach-to-grasp tasks. Maximal grasp aperture was no different between groups, but reached earlier by people with cSCI. Participants with a cSCI were less synchronous than younger and older adults but all groups used the deceleration phase for error correction to end the movement in a synchronous fashion. Overall, this study suggests that after cSCI a level of bimanual coordination is retained. While there seems to be a greater reliance on feedback to produce both the reach to grasp, we observed minimal disruption of the more impaired limb on the less impaired limb. This suggests that bimanual movements should be integrated into therapy.

  18. Influence of conspecific and heterospecific aggregation cues and alarm odors on shelter choice by syntopic spiny lobsters.

    Science.gov (United States)

    Briones-Fourzán, Patricia; Ramírez-Zaldívar, Eunice; Lozano-Alvarez, Enrique

    2008-10-01

    In spiny lobsters, conspecific scents ("aggregation cues") may mediate gregarious diurnal sheltering, but scents from injured conspecifics ("alarm odors") may elicit avoidance behavior. In laboratory experiments, individuals of two coexisting species, Panulirus guttatus (a reef-obligate) and P. argus (a temporary reef-dweller), significantly chose shelters emanating conspecific aggregation cues and responded randomly to shelters emanating heterospecific aggregation cues. However, despite evidence that the two species perceived each other's alarm odors to a similar extent, P. guttatus responded randomly to shelters emanating either conspecific or heterospecific alarm odors, whereas P. argus significantly avoided both. This differential influence of alarm odors likely reflects interspecific differences in life history, sociality, and behavior. The less social, reef-obligate P. guttatus lobsters forage close to their reef dens, into which they retract deeply upon perception of risk. This cryptic behavior may offset the need to avoid conspecific (and heterospecific) alarm odors. In contrast, avoidance of conspecific alarm odors by P. argus is consistent with its ontogenetic habitat shifts and greater sociality. Furthermore, because reef-dwelling P. argus lobsters forage across open areas away from the reef, an ability to avoid alarm odors from P. guttatus upon returning to their reef dens may increase their fitness.

  19. Application control chart concepts of designing a pre-alarm system in the nuclear power plant control room

    International Nuclear Information System (INIS)

    Hwang, S.-L.; Lin, J.-T.; Liang, G.-F.; Yau, Y.-J.; Yenn, T.-C.; Hsu, C.-C.

    2008-01-01

    This study applied the concepts of the Shewhart control chart to design a pre-alarm system for the nuclear power plant control room. As a support in detecting faults, the pre-alarm system reminded the operators of a change in the system state in its early stages. Two pre-alarm types were designed to compare with the original system, and all participants were requested to monitor each simulated system under both normal and abnormal states. The tasks for the participants included shutting down the reactor, searching for procedures, monitoring system parameters and executing secondary tasks. In each trial, the task performance, mental workload and situation awareness (SA) of the participants were measured. Results indicated that participants had lower mental workload, but equal SA, when monitoring the system with either type of pre-alarm designs, and lower alarm frequency and higher secondary task performance were obtained with the pre-alarm design. Therefore, the pre-alarm system effectively assisted the operators in monitoring tasks

  20. Advanced pulse oximeter signal processing technology compared to simple averaging. I. Effect on frequency of alarms in the operating room.

    Science.gov (United States)

    Rheineck-Leyssius, A T; Kalkman, C J

    1999-05-01

    To determine the effect of a new signal processing technique (Oxismart, Nellcor, Inc., Pleasanton, CA) on the incidence of false pulse oximeter alarms in the operating room (OR). Prospective observational study. Nonuniversity hospital. 53 ASA physical status I, II, and III consecutive patients undergoing general anesthesia with tracheal intubation. In the OR we compared the number of alarms produced by a recently developed third generation pulse oximeter (Nellcor Symphony N-3000) with Oxismart signal processing technique and a conventional pulse oximeter (Criticare 504). Three pulse oximeters were used simultaneously in each patient: a Nellcor pulse oximeter, a Criticare with the signal averaging time set at 3 seconds (Criticareaverage3s) and a similar unit with the signal averaging time set at 21 seconds (Criticareaverage21s). For each pulse oximeter, the number of false (artifact) alarms was counted. One false alarm was produced by the Nellcor (duration 55 sec) and one false alarm by the Criticareaverage21s monitor (5 sec). The incidence of false alarms was higher in Criticareaverage3s. In eight patients, Criticareaverage3s produced 20 false alarms (p signal processing compared with the Criticare monitor with the longer averaging time of 21 seconds.