WorldWideScience

Sample records for radiochromic ebt film

  1. Comparison of calibration curve of radiochromic films EBT2 and EBT

    International Nuclear Information System (INIS)

    Parra Osorio, V.; Martin-Viera Cueto, J. A.; Galan Montenegro, P.; Benitez Villegas, E. M.; Casado Villalon, F. F.; Bodineau Gil, C.

    2013-01-01

    The aim is to compare the quality of the fit to calibrate two radiochromic films batches, one model and another of EBT3 EBT2, using both experimental settings as phenomenological expression as of the calibration curve depends on the precision and accuracy of the estimate of absorbed dose. (Author)

  2. Changes of the optical characteristics of radiochromic films in the transition from EBT3 to EBT-XD films

    Science.gov (United States)

    Schoenfeld, Andreas A.; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-07-01

    A new type of radiochromic film, the EBT-XD film, has been introduced with the aim to reduce the orientation effect and the lateral response artifact occurring in the use of radiochromic films together with flatbed scanners. The task of the present study is to quantify the changes of optical characteristics involved with the transition from the well-known EBT3 films to the new EBT-XD films, using the optical bench arrangement already applied by Schoenfeld et al (2014 Phys. Med. Biol. 59 3575-97). Largely reduced polarization effects and the almost complete loss of the anisotropy of the scattered light produced in a radiation-exposed film have been observed. The Rayleigh-Debye-Gans theory is used to understand these optical changes as arising from the reduced length-to-width ratio of the LiPCDA polymer crystals in the active layer of the EBT-XD film. The effect of these changes on the flatbed scanning artifacts will be shortly addressed, but treated in more detail in a further paper.

  3. Measuring solar UV radiation with EBT radiochromic film

    International Nuclear Information System (INIS)

    Butson, Ethan T; Cheung Tsang; Yu, Peter K N; Butson, Martin J

    2010-01-01

    Ultraviolet radiation dosimetry has been performed with the use of a radiochromic film dosimeter called Gafchromic EBT for solar radiation exposure. The film changes from a clear colour to blue colour when exposed to ultraviolet radiation and results have shown that the colour change is reproducible within ±10% at 5 kJ m -2 UV exposure under various conditions of solar radiation. Parameters tested included changes in season (summer versus winter exposure), time of day, as well as sky conditions such as cloudy skies versus clear skies. As the radiochromic films' permanent colour change occurs in the visible wavelengths the film can be analysed with a desktop scanner with the most sensitive channel for analysis being the red component of the signal. Results showed that an exposure of 5 kJ m -2 (approximately 1 h exposure in full sun during summer) produced an approximate 0.28 change in the net OD when analysed in reflection mode on the desktop scanner which is significant darkening. The main advantages of this film type, and thus the new EBT2 film which has replaced EBT for measurement of UV exposure, is the visible colour change and thus easy analysis using a desktop scanner, its uniformity in response and its robust physical strength for use in outside exposure situations. (note)

  4. Radiochromic film dosimetry. Considerations on precision and accuracy for EBT2 and EBT3 type films

    Energy Technology Data Exchange (ETDEWEB)

    Dreindl, Ralf [Medical Univ. of Vienna/Vienna General Hospital (Austria). Dept. of Radiooncology; EBG MedAustron GmbH, Wiener Neustadt (Austria); Georg, Dietmar; Stock, Markus [Medical Univ. of Vienna/Vienna General Hospital (Austria). Dept. of Radiooncology; Medical Univ. of Vienna (Austria). Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology

    2014-09-01

    Gafchromic {sup registered} EBT2 film is a widely used dosimetric tool for quality assurance in radiation therapy. In 2012 EBT3 was presented as a replacement for EBT2 films. The symmetric structure of EBT3 films to reduce face-up/down dependency as well as the inclusion of a matte film surface to frustrate Newton Ring artifacts present the most prominent improvements of EBT3 films. The aim of this study was to investigate the characteristics of EBT3 films, to benchmark the films against the known EBT2-features and to evaluate the dosimetric behavior over a time period greater than 6 months. All films were irradiated to clinical photon beams (6MV, 10MV and 18MV) on an Elekta Synergy Linac equipped with a Beam Modulator MLC in solid water phantom slabs. Film digitalization was done with a flatbed transparency scanner (Type Epson Expression 1680 Pro). MATLAB {sup registered} was used for further statistical calculations and image processing. The investigations on post-irradiation darkening, film orientation, film uniformity and energy dependency resulted in negligible differences between EBT2 and EBT3 film. A minimal improvement in face-up/down dependence was found for EBT3. The matte film surface of EBT3 films turned out to be a practical feature as Newton rings could be eliminated completely. Considering long-term behavior (> 6 months) a shift of the calibration curve for EBT2 and EBT3 films due to changes in the dynamic response of the active component was observed. In conclusion, the new EBT3 film yields comparable results to its predecessor EBT2. The general advantages of radiochromic film dosimeters are completed by high film homogeneity, low energy dependence for the observed energy range and a minimized face-up/down dependence. EBT2 dosimetry-protocols can also be used for EBT3 films, but the inclusion of periodical recalibration-interval (e.g. once a quarter) is recommended for protocols of both film generations. (orig.)

  5. Radiochromic film dosimetry: considerations on precision and accuracy for EBT2 and EBT3 type films.

    Science.gov (United States)

    Dreindl, Ralf; Georg, Dietmar; Stock, Markus

    2014-05-01

    Gafchromic® EBT2 film is a widely used dosimetric tool for quality assurance in radiation therapy. In 2012 EBT3 was presented as a replacement for EBT2 films. The symmetric structure of EBT3 films to reduce face-up/down dependency as well as the inclusion of a matte film surface to frustrate Newton Ring artifacts present the most prominent improvements of EBT3 films. The aim of this study was to investigate the characteristics of EBT3 films, to benchmark the films against the known EBT2-features and to evaluate the dosimetric behavior over a time period greater than 6 months. All films were irradiated to clinical photon beams (6 MV, 10 MV and 18 MV) on an Elekta Synergy Linac equipped with a Beam Modulator MLC in solid water phantom slabs. Film digitalization was done with a flatbed transparency scanner (Type Epson Expression 1680 Pro). MATLAB® was used for further statistical calculations and image processing. The investigations on post-irradiation darkening, film orientation, film uniformity and energy dependency resulted in negligible differences between EBT2 and EBT3 film. A minimal improvement in face-up/down dependence was found for EBT3. The matte film surface of EBT3 films turned out to be a practical feature as Newton rings could be eliminated completely. Considering long-term behavior (> 6 months) a shift of the calibration curve for EBT2 and EBT3 films due to changes in the dynamic response of the active component was observed. In conclusion, the new EBT3 film yields comparable results to its predecessor EBT2. The general advantages of radiochromic film dosimeters are completed by high film homogeneity, low energy dependence for the observed energy range and a minimized face-up/down dependence. EBT2 dosimetry-protocols can also be used for EBT3 films, but the inclusion of periodical recalibration-interval (e.g. once a quarter) is recommended for protocols of both film generations. Copyright © 2013. Published by Elsevier GmbH.

  6. Radiochromic film dosimetry. Considerations on precision and accuracy for EBT2 and EBT3 type films

    International Nuclear Information System (INIS)

    Dreindl, Ralf; Georg, Dietmar; Stock, Markus; Medical Univ. of Vienna

    2014-01-01

    Gafchromic registered EBT2 film is a widely used dosimetric tool for quality assurance in radiation therapy. In 2012 EBT3 was presented as a replacement for EBT2 films. The symmetric structure of EBT3 films to reduce face-up/down dependency as well as the inclusion of a matte film surface to frustrate Newton Ring artifacts present the most prominent improvements of EBT3 films. The aim of this study was to investigate the characteristics of EBT3 films, to benchmark the films against the known EBT2-features and to evaluate the dosimetric behavior over a time period greater than 6 months. All films were irradiated to clinical photon beams (6MV, 10MV and 18MV) on an Elekta Synergy Linac equipped with a Beam Modulator MLC in solid water phantom slabs. Film digitalization was done with a flatbed transparency scanner (Type Epson Expression 1680 Pro). MATLAB registered was used for further statistical calculations and image processing. The investigations on post-irradiation darkening, film orientation, film uniformity and energy dependency resulted in negligible differences between EBT2 and EBT3 film. A minimal improvement in face-up/down dependence was found for EBT3. The matte film surface of EBT3 films turned out to be a practical feature as Newton rings could be eliminated completely. Considering long-term behavior (> 6 months) a shift of the calibration curve for EBT2 and EBT3 films due to changes in the dynamic response of the active component was observed. In conclusion, the new EBT3 film yields comparable results to its predecessor EBT2. The general advantages of radiochromic film dosimeters are completed by high film homogeneity, low energy dependence for the observed energy range and a minimized face-up/down dependence. EBT2 dosimetry-protocols can also be used for EBT3 films, but the inclusion of periodical recalibration-interval (e.g. once a quarter) is recommended for protocols of both film generations. (orig.)

  7. Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams

    Science.gov (United States)

    Khachonkham, Suphalak; Dreindl, Ralf; Heilemann, Gerd; Lechner, Wolfgang; Fuchs, Hermann; Palmans, Hugo; Georg, Dietmar; Kuess, Peter

    2018-03-01

    Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure as the EBT3 film but has a slightly different composition and a thinner active layer. This study benchmarks the EBT-XD against EBT3 film for 6 MV and 10 MV photon beams, as well as for 97.4 MeV and 148.2 MeV proton beams and 15-100 kV x-rays. Dosimetric and film reading characteristics, such as post irradiation darkening, film orientation effect, lateral response artifact (LRA), film sensitivity, energy and beam quality dependency were investigated. Furthermore, quenching effects in the Bragg peak were investigated for a single proton beam energy for both film types, in addition measurements were performed in a spread-out Bragg peak. EBT-XD films showed the same characteristic on film darkening as EBT3. The effects between portrait and landscape orientation were reduced by 3.1% (in pixel value) for EBT-XD compared to EBT3 at a dose of 2000 cGy. The LRA is reduced for EBT-XD films for all investigated dose ranges. The sensitivity of EBT-XD films is superior to EBT3 for doses higher than 500 cGy. In addition, EBT-XD showed a similar dosimetric response for photon and proton irradiation with low energy and beam quality dependency. A quenching effect of 10% was found for both film types. The slight decrease in the thickness of the active layer and different composition configuration of EBT-XD resulted in a reduced film orientation effect and LRA, as well as a sensitivity increase in high-dose regions for both photon and proton beams. Overall, the EBT-XD film improved regarding film reading characteristics and showed advantages in the high-dose region for photon and proton beams.

  8. Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams.

    Science.gov (United States)

    Khachonkham, Suphalak; Dreindl, Ralf; Heilemann, Gerd; Lechner, Wolfgang; Fuchs, Hermann; Palmans, Hugo; Georg, Dietmar; Kuess, Peter

    2018-03-15

    Recently, a new type of radiochromic film, the EBT-XD film, has been introduced for high dose radiotherapy. The EBT-XD film contains the same structure as the EBT3 film but has a slightly different composition and a thinner active layer. This study benchmarks the EBT-XD against EBT3 film for 6 MV and 10 MV photon beams, as well as for 97.4 MeV and 148.2 MeV proton beams and 15-100 kV x-rays. Dosimetric and film reading characteristics, such as post irradiation darkening, film orientation effect, lateral response artifact (LRA), film sensitivity, energy and beam quality dependency were investigated. Furthermore, quenching effects in the Bragg peak were investigated for a single proton beam energy for both film types, in addition measurements were performed in a spread-out Bragg peak. EBT-XD films showed the same characteristic on film darkening as EBT3. The effects between portrait and landscape orientation were reduced by 3.1% (in pixel value) for EBT-XD compared to EBT3 at a dose of 2000 cGy. The LRA is reduced for EBT-XD films for all investigated dose ranges. The sensitivity of EBT-XD films is superior to EBT3 for doses higher than 500 cGy. In addition, EBT-XD showed a similar dosimetric response for photon and proton irradiation with low energy and beam quality dependency. A quenching effect of 10% was found for both film types. The slight decrease in the thickness of the active layer and different composition configuration of EBT-XD resulted in a reduced film orientation effect and LRA, as well as a sensitivity increase in high-dose regions for both photon and proton beams. Overall, the EBT-XD film improved regarding film reading characteristics and showed advantages in the high-dose region for photon and proton beams.

  9. Analysis of the response dependence of Ebt3 radiochromic film with energy, dose rate, wavelength, scanning mode and humidity

    International Nuclear Information System (INIS)

    Leon M, E. Y.; Camacho L, M. A.; Herrera G, J. A.; Garcia G, O. A.; Villarreal B, J. E.

    2016-10-01

    With the development of new modalities in radiotherapy treatments, the use of radiochromic films has increased considerably. Because the characteristics that presented, they are suitable for quality control and dose measurement. In this work and analysis of the dependence of the response of Ebt3 radiochromic films with energy, dose rate, wavelength, scan mode and humidity, for a dose range of 0-70 Gy is presented. According to the results, the response of Ebt3 radiochromic films has low dependence on energy, dose rate, scan mode and humidity. However, the sensitivity of the response Ebt3 radiochromic films has a high dependence on the wavelength of the optical system used for reading. (Author)

  10. Comparison of calibration curve of radiochromic films EBT2 and EBT; Comparacion de la curva de calibracion de las peliculas radiocromica EBT2 y EBT

    Energy Technology Data Exchange (ETDEWEB)

    Parra Osorio, V.; Martin-Viera Cueto, J. A.; Galan Montenegro, P.; Benitez Villegas, E. M.; Casado Villalon, F. F.; Bodineau Gil, C.

    2013-07-01

    The aim is to compare the quality of the fit to calibrate two radiochromic films batches, one model and another of EBT3 EBT2, using both experimental settings as phenomenological expression as of the calibration curve depends on the precision and accuracy of the estimate of absorbed dose. (Author)

  11. Model changes EBT2 radiochromic film from its predecessor EBT; Cambios del modelo EBT2 de peliculas radiocromicas respecto a su predecesora EBT

    Energy Technology Data Exchange (ETDEWEB)

    Carrasco Herrera, M. A.; Perucha Ortega, M.; Baeza Trujillo, M.; Luis Simon, F. J.; Herrador Cordoba, M.

    2011-07-01

    The model EBT2 Gafchromic radiochromic film is significantly different from its predecessor EBT. The presence of a yellow pigment in the active layer results in a reduced sensitivity to ambient light, on the other hand, this pigment can apply a correction to the measured signal to compensate for changes due to differences in the thickness of the active layer, there by improving a priori, the homogeneity of the response of the film. Another new feature they present is the lack of symmetry of the layers that make up the film, leading to the emergence of a new unit of the pixel value obtained with the orientation of the film to scan: heads or tails. (Author)

  12. Characterization of the proton beam from an IBA Cyclone 18/9 with radiochromic film EBT2

    Energy Technology Data Exchange (ETDEWEB)

    Sansaloni, F.; Lagares, J. I.; Arce, P.; Llop, J.; Perez, J. M. [Medical Applications Unit, Technology Department, CIEMAT, Madrid (Spain); Radiochemistry Department, Molecular Imaging unit, CIC-biomaGUNE, San Sebastian (Spain); Technology Department, CIEMAT (Spain)

    2012-12-19

    The use of radiochromic films is widespread in different areas of medical physics like radiotherapy and hadrontherapy; however, radiochromic films have been scarcely used in the characterization of proton or deuteron beams generated in biomedical cyclotrons. In this paper the radiochromic film EBT2 was used to study the beam size and the proton beam energy of an IBA Cyclone 18/9 cyclotron. The results indicate that the beam size can be easily measured at a very low expense; however, an accurate determination of the beam energy might require the implementation of certain experimental improvements.

  13. Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT GAFCHROMIC[reg] film

    International Nuclear Information System (INIS)

    Lynch, Bart D.; Kozelka, Jakub; Ranade, Manisha K.; Li, Jonathan G.; Simon, William E.; Dempsey, James F.

    2006-01-01

    In this study, we present three significant artifacts that have the potential to negatively impact the accuracy and precision of film dosimetry measurements made using GAFCHROMIC[reg] EBT radiochromic film when read out with CCD flatbed scanners. Films were scanned using three commonly employed instruments: a Macbeth TD932 spot densitometer, an Epson Expression 1680 CCD array scanner, and a Microtek ScanMaker i900 CCD array scanner. For the two scanners we assessed the variation in optical density (OD) of GAFCHROMIC EBT film with scanning bed position, angular rotation of the film with respect to the scan line direction, and temperature inside the scanner due to repeated scanning. Scanning uniform radiochromic films demonstrated a distinct bowing effect in profiles in the direction of the CCD array with a nonuniformity of up to 17%. Profiles along a direction orthogonal to the CCD array demonstrated a 7% variation. A strong angular dependence was found in measurements made with the flatbed scanners; the effect could not be reproduced with the spot densitometer. An IMRT quality assurance film was scanned twice rotating the film 90 deg. between the scans. For films scanned on the Epson scanner, up to 12% variation was observed in unirradiated EBT films rotated between 0 deg. and 90 deg. , which decreased to approximately 8% for EBT films irradiated to 300 cGy. Variations of up to 80% were observed for films scanned with the Microtek scanner. The scanners were found to significantly increase the film temperature with repeated scanning. Film temperature between 18 and 33 deg. C caused OD changes of approximately 7%. Considering these effects, we recommend adherence to a strict scanning protocol that includes: maintaining the orientation of films scanned on flatbed scanners, limiting scanning to the central portion of the scanner bed, and limiting the number of consecutive scans to minimize changes in OD caused by film heating

  14. Calibration in water films GAFCHROMIC EBT radiochromic-2. Effects of Drying

    International Nuclear Information System (INIS)

    Herrero, C.; Perez-Alija, J.; Alaman, C.; Almansa, J.; Vilches, M.

    2011-01-01

    Recent studies [1) show that immersion in water GAFCHROMIC R EBT radiochromic film-2 has on these two effects: a progressive diffusion of water across its borders (fact which tells us the manufacturer and the effect of which we can avoid choosing a proper analysis region) and a uniform moisture through the polyester film, producing a variation in the measurement of optical density (OD) of the film. The latter effect is negligible for immersion times of less than 30 minutes. We study the effect of hydration and drying in these films when they have been submerged for 24 hours.

  15. Evaluation of a novel triple-channel radiochromic film analysis procedure using EBT2.

    Science.gov (United States)

    van Hoof, Stefan J; Granton, Patrick V; Landry, Guillaume; Podesta, Mark; Verhaegen, Frank

    2012-07-07

    A novel approach to read out radiochromic film was introduced recently by the manufacturer of GafChromic film. In this study, the performance of this triple-channel film dosimetry method was compared against the conventional single-red-channel film dosimetry procedure, with and without inclusion of a pre-irradiation (pre-IR) film scan, using EBT2 film and kilo- and megavoltage photon beams up to 10 Gy. When considering regions of interest averaged doses, the triple-channel method and both single-channel methods produced equivalent results. Absolute dose discrepancies between the triple-channel method, both single-channel methods and the treatment planning system calculated dose values, were no larger than 5 cGy for dose levels up to 2.2 Gy. Signal to noise in triple-channel dose images was found to be similar to signal to noise in single-channel dose images. The accuracy of resulting dose images from the triple- and single-channel methods with inclusion of pre-IR film scan was found to be similar. Results of a comparison of EBT2 data from a kilovoltage depth dose experiment to corresponding Monte Carlo depth dose data produced dose discrepancies of 9.5 ± 12 cGy and 7.6 ± 6 cGy for the single-channel method with inclusion of a pre-IR film scan and the triple-channel method, respectively. EBT2 showed to be energy sensitive at low kilovoltage energies with response differences of 11.9% and 15.6% in the red channel at 2 Gy between 50-225 kVp and 80-225 kVp photon spectra, respectively. We observed that the triple-channel method resulted in non-uniformity corrections of ±1% and consistency values of 0-3 cGy for the batches and dose levels studied. Results of this study indicate that the triple-channel radiochromic film read-out method performs at least as well as the single-channel method with inclusion of a pre-IR film scan, reduces film non-uniformity and saves time with elimination of a pre-IR film scan.

  16. Evaluation of a novel triple-channel radiochromic film analysis procedure using EBT2

    International Nuclear Information System (INIS)

    Van Hoof, Stefan J; Granton, Patrick V; Landry, Guillaume; Podesta, Mark; Verhaegen, Frank

    2012-01-01

    A novel approach to read out radiochromic film was introduced recently by the manufacturer of GafChromic film. In this study, the performance of this triple-channel film dosimetry method was compared against the conventional single-red-channel film dosimetry procedure, with and without inclusion of a pre-irradiation (pre-IR) film scan, using EBT2 film and kilo- and megavoltage photon beams up to 10 Gy. When considering regions of interest averaged doses, the triple-channel method and both single-channel methods produced equivalent results. Absolute dose discrepancies between the triple-channel method, both single-channel methods and the treatment planning system calculated dose values, were no larger than 5 cGy for dose levels up to 2.2 Gy. Signal to noise in triple-channel dose images was found to be similar to signal to noise in single-channel dose images. The accuracy of resulting dose images from the triple- and single-channel methods with inclusion of pre-IR film scan was found to be similar. Results of a comparison of EBT2 data from a kilovoltage depth dose experiment to corresponding Monte Carlo depth dose data produced dose discrepancies of 9.5 ± 12 cGy and 7.6 ± 6 cGy for the single-channel method with inclusion of a pre-IR film scan and the triple-channel method, respectively. EBT2 showed to be energy sensitive at low kilovoltage energies with response differences of 11.9% and 15.6% in the red channel at 2 Gy between 50–225 kVp and 80–225 kVp photon spectra, respectively. We observed that the triple-channel method resulted in non-uniformity corrections of ±1% and consistency values of 0–3 cGy for the batches and dose levels studied. Results of this study indicate that the triple-channel radiochromic film read-out method performs at least as well as the single-channel method with inclusion of a pre-IR film scan, reduces film non-uniformity and saves time with elimination of a pre-IR film scan. (paper)

  17. Analysis of the response dependence of Ebt3 radiochromic film with energy, dose rate, wavelength, scanning mode and humidity; Analisis de la dependencia de la respuesta de la pelicula radiocromica EBT3 con la energia, tasa de dosis, longitud de onda, modo de escaneo y con la humedad

    Energy Technology Data Exchange (ETDEWEB)

    Leon M, E. Y.; Camacho L, M. A. [Universidad Autonoma del Estado de Mexico, Facultad de Medicina, Laboratorio de Fotomedicina, Biofotonica y Espectroscopia Laser de Pulsos Ultracortos, Jesus Carranza y Paseo Tollocan s/n, 50120 Toluca, Estado de Mexico (Mexico); Herrera G, J. A.; Garcia G, O. A. [Instituto Nacional de Neurologia y Neurocirugia, Laboratorio de Fisica Medica y Unidad de Radiocirugia, 14269 Ciudad de Mexico (Mexico); Villarreal B, J. E., E-mail: yaz_3333@hotmail.com [University of Calgary, Department of Oncology, Tom Baker Cancer Centre, 1331 29th street NW Calgary, Alberta T2N 4N2 (Canada)

    2016-10-15

    With the development of new modalities in radiotherapy treatments, the use of radiochromic films has increased considerably. Because the characteristics that presented, they are suitable for quality control and dose measurement. In this work and analysis of the dependence of the response of Ebt3 radiochromic films with energy, dose rate, wavelength, scan mode and humidity, for a dose range of 0-70 Gy is presented. According to the results, the response of Ebt3 radiochromic films has low dependence on energy, dose rate, scan mode and humidity. However, the sensitivity of the response Ebt3 radiochromic films has a high dependence on the wavelength of the optical system used for reading. (Author)

  18. Multi-dimensional dosimetric verification of stereotactic radiotherapy for uveal melanoma using radiochromic EBT film

    International Nuclear Information System (INIS)

    Sturtewagen, E.; Fuss, M.; Georg, D.; Paelinck, L.; Wagter, C. de

    2008-01-01

    Since 1997, linac based stereotactic radiotherapy (SRT) of uveal melanoma has been continuously developed at the Department of Radiotherapy, Medical University Vienna. The aim of the present study was (i) to test a new type of radiochromic film (Gafchromic EBT) for dosimetric verification of class solutions for these treatments and (ii) to verify treatment plan acceptance criteria, which are based on gamma values statisitcs. An EPSON Expression 1680 Pro flat bed scanner was utilized for film reading. To establish a calibration curve, films were cut in squares of 2 x 2 cm 2 , positioned at 5 cm depth in a solid water phantom and were irradiated with different dose levels (0.5 and 5 Gy) in a 5 x 5 cm 2 field at 6 MV. A previously developed solid phantom (polystyrene) was used with overall dimensions corresponding to an average human head. EBT films were placed at four different depths (10, 20, 25 and 30 mm) and all films were irradiated simultaneously. Four different treatment plans were verified that resemble typical clinical situations. These plans differed in irradiation technique (conformal mMLC or circular arc SRT) and in tumour size (PTV of 1 or 2.5 cm 3 ). In-house developed software was applied to calculate gamma (γ) index values and to perform several statistical operations (e.g. γ-area histograms). At depths of 10 mm γ 1% (γ-value where 1% of the points have an equal or higher value in the region of interest) were between 1-3 and maximum γ > 1 (% of γ-values > 1 in the region of interest) areas were almost 30%. At larger depths, i.e. more close to the isocenter, γ 1% was > 1 areas were mostly < 5%. Average γ values were about 0.5. Besides the compromised accuracy in the buildup region, previously defined IMRT acceptance criteria [Stock et al., Phys. Med. Biol. 50 (2005) 399-411] could be applied as well to SRT. Radiochromic EBT films, in combination with a flat-bed scanner, were found to be an ideal multidimensional dosimetric tool for treatment

  19. Dosimetric comparison between a planning system and Radiochromic-EBT2 films in surface brachytherapy treatments of high rate

    International Nuclear Information System (INIS)

    Gutierrez Ramos, S. M.; Carrasco Herrera, M. a.; Vicent, D.; Rodriguez, C.; Herrador, M.

    2013-01-01

    The objective of this work is to study a situation in which, the accuracy of the calculation of the planner may be limited, superficial brachytherapy treatment. It has relative to the dose obtained with the planner with that obtained with film radiochromic EBT2. (Author)

  20. A preliminary investigation of the EBT2 radiochromic films response to low energy fast neutrons

    Energy Technology Data Exchange (ETDEWEB)

    Aydarous, Abdulkadir, E-mail: Aydarous@gmail.com [Physics Department, Faculty of Science, Taif University, Al-Hawiah, Taif, PO Box 888 (Saudi Arabia); Faculty of Energy Systems and Nuclear Science, University of Ontario Institute of Technology, 2000 Simcoe St North, Oshawa, ON, L1H 7K4 (Canada); Aslam [Faculty of Energy Systems and Nuclear Science, University of Ontario Institute of Technology, 2000 Simcoe St North, Oshawa, ON, L1H 7K4 (Canada); Waker, Anthony [Faculty of Energy Systems and Nuclear Science, University of Ontario Institute of Technology, 2000 Simcoe St North, Oshawa, ON, L1H 7K4 (Canada); Department of Medical Physics and Applied Radiation Sciences, McMaster University, Hamilton, ON, L8S 4M1 (Canada)

    2012-07-15

    EBT2 radiochromic films were used to study the relative dose distribution of the neutron field. The correlation between the beam current and the optical density showed good linear dependence with a correlation coefficient exceeding 98%. At any given beam energy, neutron dose rates can be changed by a factor of 40 without changing the neutron spectrum. This result is consistent with what was found by the Tissue Equivalent Proportional Counter measurements. The uniformity of the neutron field was inspected by the optical density profile of the exposed film. - Highlights: Black-Right-Pointing-Pointer Developing a 2D image for neutron field. Black-Right-Pointing-Pointer Investigation of EBT2 sensitivity to neutrons. Black-Right-Pointing-Pointer Studying the effect of irradiation parameters (beam energy and beam current) to the measured optical density.

  1. Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy

    International Nuclear Information System (INIS)

    Richley, L; John, A C; Coomber, H; Fletcher, S

    2010-01-01

    A new radiochromic film, the yellow Gafchromic EBT2, has been marketed as a drop-in replacement for the discontinued blue EBT film. In order to verify the manufacturer's claims prior to clinical use, EBT2 was characterized in transmission, and the less commonly used, reflection modes with an Epson Expression 10000XL A3 flatbed scanner. The red channel was confirmed to provide the greatest sensitivity and was used for all measurements. The post-irradiation darkening of the film was investigated, and the relative response was found to be dose dependent with higher doses stabilizing earlier than lower doses. After 13 h all dose levels had stabilized to within 1% of their value at 24 h. Uniformity of irradiated EBT2 films was within 0.8% and 1.2% (2SD of signal), in reflection and transmission modes, respectively. The light scattering effect, arising from the structure and thickness of EBT2, was found to give rise to an apparent scanner non-uniformity of up to 5.5% in signal. In reflection mode, differences of up to 1.2% were found between the signal obtained from a small film fragment (5 x 5 cm 2 ) and the signal obtained from the same fragment bordered by extra film. Further work is needed to determine the origin of this effect, as there will be implications for reflection dosimetry of intensity modulated fields; reflection mode cannot yet be regarded as a viable alternative to transmission mode. Our results suggest that EBT2 film is a valid alternative, rather than a direct replacement for EBT film.

  2. Magnetic fields are causing small, but significant changes of the radiochromic EBT3 film response to 6 MV photons.

    Science.gov (United States)

    Delfs, Björn; Schoenfeld, Andreas A; Poppinga, Daniela; Kapsch, Ralf-Peter; Jiang, Ping; Harder, Dietrich; Poppe, Björn; Looe, Hui Khee

    2018-01-31

    The optical density (OD) of EBT3 radiochromic films (Ashland Specialty Ingredients, Bridgewater, NJ, USA) exposed to absorbed doses to water up to D  =  20 Gy in magnetic fields of B  =  0.35 and 1.42 T was measured in the three colour channels of an Epson Expression 10000XL flatbed scanner. A 7 cm wide water phantom with fixed film holder was placed between the pole shoes of a constant-current electromagnet with variable field strength and was irradiated by a 6 MV photon beam whose axis was directed at right angles with the field lines. The doses at the film position at water depth 5 cm were measured with a calibrated ionization chamber when the magnet was switched off and were converted to the doses in presence of the magnetic field via the monitor units and by a Monte Carlo-calculated correction accounting for the slight change of the depth dose curves in magnetic fields. In the presence of the 0.35 and 1.42 T fields small negative changes of the OD values at given absorbed doses to water occurred and just significantly exceeded the uncertainty margin given by the stochastic and the uncorrected systematic deviations. This change can be described by a  +2.1% change of the dose values needed to produce a given optical density in the presence of a 1.42 T field. The thereby modified OD versus D function remained unchanged irrespective of whether the original short film side-the preference direction of the monomer crystals of the film-was directed parallel or orthogonal to the magnetic field. The 'orientation effect', the difference between the optical densities measured in the 'portrait' or 'landscape' film positions on the scanner bed caused by the reflection of polarised light in the scanner's mirror system, remained unaltered after EBT3 film exposure in magnetic fields. An independent optical bench investigation of EBT3 films exposed to doses of 10 and 20 Gy at 0.35 and 1.42 T showed that the direction of the electric vector of polarised

  3. Magnetic fields are causing small, but significant changes of the radiochromic EBT3 film response to 6 MV photons

    Science.gov (United States)

    Delfs, Björn; Schoenfeld, Andreas A.; Poppinga, Daniela; Kapsch, Ralf-Peter; Jiang, Ping; Harder, Dietrich; Poppe, Björn; Khee Looe, Hui

    2018-02-01

    The optical density (OD) of EBT3 radiochromic films (Ashland Specialty Ingredients, Bridgewater, NJ, USA) exposed to absorbed doses to water up to D  =  20 Gy in magnetic fields of B  =  0.35 and 1.42 T was measured in the three colour channels of an Epson Expression 10000XL flatbed scanner. A 7 cm wide water phantom with fixed film holder was placed between the pole shoes of a constant-current electromagnet with variable field strength and was irradiated by a 6 MV photon beam whose axis was directed at right angles with the field lines. The doses at the film position at water depth 5 cm were measured with a calibrated ionization chamber when the magnet was switched off and were converted to the doses in presence of the magnetic field via the monitor units and by a Monte Carlo-calculated correction accounting for the slight change of the depth dose curves in magnetic fields. In the presence of the 0.35 and 1.42 T fields small negative changes of the OD values at given absorbed doses to water occurred and just significantly exceeded the uncertainty margin given by the stochastic and the uncorrected systematic deviations. This change can be described by a  +2.1% change of the dose values needed to produce a given optical density in the presence of a 1.42 T field. The thereby modified OD versus D function remained unchanged irrespective of whether the original short film side—the preference direction of the monomer crystals of the film—was directed parallel or orthogonal to the magnetic field. The ‘orientation effect’, the difference between the optical densities measured in the ‘portrait’ or ‘landscape’ film positions on the scanner bed caused by the reflection of polarised light in the scanner’s mirror system, remained unaltered after EBT3 film exposure in magnetic fields. An independent optical bench investigation of EBT3 films exposed to doses of 10 and 20 Gy at 0.35 and 1.42 T showed that the direction of the electric

  4. Calibration in water films GAFCHROMIC EBT radiochromic-2. Effects of Drying; Calibracion en agua de las peliculas radiocromicas Gafchromic EBT-2. Efectos de Secado

    Energy Technology Data Exchange (ETDEWEB)

    Herrero, C.; Perez-Alija, J.; Alaman, C.; Almansa, J.; Vilches, M.

    2011-07-01

    Recent studies [1] show that immersion in water GAFCHROMIC EBT radiochromic film-2 has on these two effects: a progressive diffusion of water across its borders (fact which tells us the manufacturer and the effect of which we can avoid choosing a proper analysis region) and a uniform moisture through the polyester film, producing a variation in the measurement of optical density (OD) of the film. The latter effect is negligible for immersion times of less than 30 minutes. We study the effect of hydration and drying in these films when they have been submerged for 24 hours.

  5. Comparison of the radiochromic EBT2 responses for 4MV LINACs in calibration processes

    Energy Technology Data Exchange (ETDEWEB)

    Silva, Sabrina D.; Castro, André L.S.; Mendes, Thais M.; Campos, Tarcisio P.R. [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Dept. de Energia Nuclear; Nogueira, Luciana B., E-mail: sadonatosilva@hotmail.com, E-mail: radioterapia.andre@gmail.com, E-mail: thaismelomendes@yahoo.com.br, E-mail: tprcampos@yahoo.com.br, E-mail: Lucibn19@yahoo.com.br [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Anatomia e Imagem

    2017-07-01

    Background: Actually, cancer has gained a larger dimension and become a global public health problem. Radiotherapy (RT) is a neoplasia treatment and RT linear accelerators must undergo a strict dose quality control. Pure liquid water or solid water phantoms can be used with this intuit. In recent years, radiochromic films with equivalent tissue composition have been widely used as dosimeters in the medical field. Objective: the proposal was to analyze two distinct radiochromic film responses in water and solid water phantoms, in a distinct LINAC spectra of 4MV beam. Methods: Solid water phantom and EBT2 Radiochromic films were set. Films were exposed to a set of absorbed doses established by distinct monitor units (MU) specified in each RT-center. Mathematical relations between the degree of red-intensity from digitized films and the absorbed dose for both spectra were established. The coefficients of the polynomial function of the calibration curve were determined from the Origin software. The uncertainty of both processes was analyzed. The efficiency of the two calibration processes was set up. Results: The adjustment of the calibration curve provided the coefficients of the second-order equation that relates the dose absorbed with the optical density of the film. The uncertainty regarding the calibration performed in water and solid water and the dose-error accuracy are in agreement with the literature. Conclusion: Both water or solid water were effective in calibration and can be used in routines of quality-control measurements. The results show that EBT2-radiochromic films is suitable to for dose-calibration in RT. (author)

  6. Characterization of radiochromic films EBT3 by means of the scanner Vidar dosimetry Pro Red and Epson 10000-XL use

    International Nuclear Information System (INIS)

    Medina, L.; Garrigo, E.; Venencia, D.; Adrada, A.; Filipuzzi, M.

    2014-08-01

    The Radiochromic film have become an attractive tool for verification of dose distributions in IMRT because these have high spatial resolution film, are near water equivalent and not require revealed, A critical aspect of the use of these film is used for digitizing scanner The purpose of this paper is to characterize EBT3 using two types of scanner. Were employed The Radiochromic film EBT3, was used photon beam 6 MV generated by a linear accelerator Siemens Primus, he films were irradiated at a dose range between 0 Gy a 9 Gy. The stabilization time after irradiation was 24 hours. The films were digitized with a scanner Epson 10000-XL y el VIDAR Dosimetry Pro Red. We used the software for construction of the calibration curve. The resolution of each dosimetry system was analyzed through the results of the spatial response function by analyzing a step pattern. The Epson scanner is most sensitive to the red channel. This is less than that obtained with the Vidar. The Vidar scanner spatial response profiles disturbs not opposed to Epson analyzed. The calibration curves for both dosimetry systems can be employed. However, the sensitivity and repeatability of the system is better than Red Vidar Epson 10000-XL. (author)

  7. Characteristics of the radiochromic film Gafchromictm EBT3 model for use in brachytherapy

    International Nuclear Information System (INIS)

    Luvizotto, Jessica

    2015-01-01

    Brachytherapy is a radiotherapy treatment modality using radioactive sealed sources within walking distance of the tumor, reducing the risk of applying an unwanted dose to adjacent healthy tissues. For brachytherapy is reliable, it is necessary to establish a dosimetric practices program aimed at determining the optimal dose of radiation for this radiotherapy practice. This paper presents the application of two methodologies for the dosimetry using radiochromic films. Experimental measurements were performed with EBT3 films in phantoms consisting of homogeneous and heterogeneous material (lung, bone and soft tissue) built especially for dose measurements in brachytherapy. The processing and analysis of the resulting images of the experimental procedure were performed with ImageJ software and MATLAB. The results were evaluated from comparisons dose of experimental measurements and simulations obtained by the Monte Carlo method. (author)

  8. Dose and absorption spectra response of EBT2 Gafchromic film to high energy X-rays

    International Nuclear Information System (INIS)

    Butson, M.J.; Cheung, T.; Yu, P.K.N.; Alnawaf, H.

    2009-01-01

    Full text: With new advancements in radiochromic film designs and sensitivity to suit different niche applications, EBT2 is the latest offering for the megavoltage radiotherapy market. New construction specifications including different physical construction and the use of a yellow coloured dye has provided the next generation radiochromic film for therapy applications. The film utilises the same active chemical for radiation measurement as its predecessor, EBT Gafchromic. Measurements have been performed using photo spectrometers to analyse the absorption spectra properties of this new EBT2 Gafchromic, radiochromic film. Results have shown that whilst the physical coloration or absorption spectra of the film, which turns yellow to green as compared to EBT film, (clear to blue) is significantly different due to the added yellow dye, the net change in absorption spectra properties for EBT2 are similar to the original EBT film. Absorption peaks are still located at 636 n m and 585 n m positions. A net optical density change of 0.590 ± 0.020 (2SD) for a 1 Gy radiation absorbed dose using 6 MV x-rays when measured at the 636 n m absorption peak was found. This is compared to 0.602 ± 0.025 (2SD) for the original EBT film (2005 Batch) and 0.557 ± 0.027 (2009 Batch) at the same absorption peak. The yellow dye and the new coating material produce a significantly different visible absorption spectra results for the EBT2 film compared to EBT at wavelengths especially below approximately 550 n m. At wavelengths above 550 n m differences in absolute OD are seen however, when dose analysis is performed at wavelengths above 550 n m using net optical density changes, no significant variations are seen. If comparing results of the late production EBT to new production EBT2 film, net optical density variations of approximately 10 % to 15 % are seen. As all new film batches should be calibrated for sensitivity upon arrival this should not be of concern.

  9. Characteristics of the radiochromic film Gafchromictm EBT3 model for use in brachytherapy; Caracterizacao do filme radiocromico Gafchromictm modelo EBT3 para uso em braquiterapia

    Energy Technology Data Exchange (ETDEWEB)

    Luvizotto, Jessica

    2015-07-01

    Brachytherapy is a radiotherapy treatment modality using radioactive sealed sources within walking distance of the tumor, reducing the risk of applying an unwanted dose to adjacent healthy tissues. For brachytherapy is reliable, it is necessary to establish a dosimetric practices program aimed at determining the optimal dose of radiation for this radiotherapy practice. This paper presents the application of two methodologies for the dosimetry using radiochromic films. Experimental measurements were performed with EBT3 films in phantoms consisting of homogeneous and heterogeneous material (lung, bone and soft tissue) built especially for dose measurements in brachytherapy. The processing and analysis of the resulting images of the experimental procedure were performed with ImageJ software and MATLAB. The results were evaluated from comparisons dose of experimental measurements and simulations obtained by the Monte Carlo method. (author)

  10. Technical Note: Homogeneity of Gafchromic EBT2 film

    International Nuclear Information System (INIS)

    Hartmann, Bernadette; Martisikova, Maria; Jaekel, Oliver

    2010-01-01

    Purpose: The self-developing Gafchromic EBT film is a radiochromic film, widely used for relative photon dosimetry. Recently, the manufacturer has replaced the well-investigated EBT film by the new Gafchromic EBT2 film. It has the same sensitive component and, in addition, it contains a yellow marker dye in order to protect the film against ambient light exposure and to serve as a base for corrections of small differences in film response. Furthermore, the configuration of the film layers as well as the binder material have been changed in comparison to the EBT film. When investigating the properties of EBT2 film, all characteristics were found to be similar to those of EBT film, except for the film response homogeneity. Thus, in this article special focus was put on examining the homogeneity of EBT2 film. Methods: A scan protocol established for EBT film and published previously was used. The uniformity of the film coloration was investigated for unirradiated and irradiated EBT2 film sheets. The dose response of EBT2 film was measured and the influence of film inhomogeneities on dose determination was evaluated. Results: Inhomogeneities in pixel values of up to ±3.7% within one film were detected. The relative inhomogeneities were found to be approximately independent of the dose. Nonuniformities of the film response lead to uncertainties in dose determination of ±8.7% at 1 Gy. When using net optical densities for dose calibration, uncertainties in dose determination amount to more than ±6%. Conclusions: EBT2 films from the lot investigated in this study show response inhomogeneities, which lead to uncertainties in dose determination exceeding the commonly accepted tolerance levels. It is important to test further EBT2 lots regarding homogeneity before using the film in clinical routine.

  11. Characterization of radiochromic films EBT3 by means of the scanner Vidar dosimetry Pro Red and Epson 10000-XL use; Caracterizacion de films radiocromicos EBT3 mediante el uso de scanner Vidar dosimetry Pro Red y Epson 10000-XL

    Energy Technology Data Exchange (ETDEWEB)

    Medina, L.; Garrigo, E.; Venencia, D.; Adrada, A.; Filipuzzi, M., E-mail: fisicamedina11@gmail.com [Instituto Privado de Radioterapia, Obispo Oro 423, X5000BFI Cordoba (Argentina)

    2014-08-15

    The Radiochromic film have become an attractive tool for verification of dose distributions in IMRT because these have high spatial resolution film, are near water equivalent and not require revealed, A critical aspect of the use of these film is used for digitizing scanner The purpose of this paper is to characterize EBT3 using two types of scanner. Were employed The Radiochromic film EBT3, was used photon beam 6 MV generated by a linear accelerator Siemens Primus, he films were irradiated at a dose range between 0 Gy a 9 Gy. The stabilization time after irradiation was 24 hours. The films were digitized with a scanner Epson 10000-XL y el VIDAR Dosimetry Pro Red. We used the software for construction of the calibration curve. The resolution of each dosimetry system was analyzed through the results of the spatial response function by analyzing a step pattern. The Epson scanner is most sensitive to the red channel. This is less than that obtained with the Vidar. The Vidar scanner spatial response profiles disturbs not opposed to Epson analyzed. The calibration curves for both dosimetry systems can be employed. However, the sensitivity and repeatability of the system is better than Red Vidar Epson 10000-XL. (author)

  12. Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification

    Science.gov (United States)

    Palmer, Antony L.; Dimitriadis, Alexis; Nisbet, Andrew; Clark, Catharine H.

    2015-11-01

    There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200-2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison.

  13. Evaluation of Gafchromic EBT-XD film, with comparison to EBT3 film, and application in high dose radiotherapy verification

    International Nuclear Information System (INIS)

    Palmer, Antony L; Dimitriadis, Alexis; Nisbet, Andrew; Clark, Catharine H

    2015-01-01

    There is renewed interest in film dosimetry for the verification of dose delivery of complex treatments, particularly small fields, compared to treatment planning system calculations. A new radiochromic film, Gafchromic EBT-XD, is available for high-dose treatment verification and we present the first published evaluation of its use. We evaluate the new film for MV photon dosimetry, including calibration curves, performance with single- and triple-channel dosimetry, and comparison to existing EBT3 film. In the verification of a typical 25 Gy stereotactic radiotherapy (SRS) treatment, compared to TPS planned dose distribution, excellent agreement was seen with EBT-XD using triple-channel dosimetry, in isodose overlay, maximum 1.0 mm difference over 200–2400 cGy, and gamma evaluation, mean passing rate 97% at 3% locally-normalised, 1.5 mm criteria. In comparison to EBT3, EBT-XD gave improved evaluation results for the SRS-plan, had improved calibration curve gradients at high doses, and had reduced lateral scanner effect. The dimensions of the two films are identical. The optical density of EBT-XD is lower than EBT3 for the same dose. The effective atomic number for both may be considered water-equivalent in MV radiotherapy. We have validated the use of EBT-XD for high-dose, small-field radiotherapy, for routine QC and a forthcoming multi-centre SRS dosimetry intercomparison. (paper)

  14. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance

    International Nuclear Information System (INIS)

    Fuss, Martina; Sturtewagen, Eva; Wagter, Carlos De; Georg, Dietmar

    2007-01-01

    The suitability of radiochromic EBT film was studied for high-precision clinical quality assurance (QA) by identifying the dose response for a wide range of irradiation parameters typically modified in highly-conformal treatment techniques. In addition, uncertainties associated with varying irradiation conditions were determined. EBT can be used for dose assessment of absorbed dose levels as well as relative dosimetry when compared to absolute absorbed dose calibrated using ionization chamber results. For comparison, a silver halide film (Kodak EDR-2) representing the current standard in film dosimetry was included. As an initial step a measurement protocol yielding accurate and precise results was established for a flatbed transparency scanner (Epson Expression 1680 Pro) that was utilized as a film reading instrument. The light transmission measured by the scanner was found to depend on the position of the film on the scanner plate. For three film pieces irradiated with doses of 0 Gy, ∼1 Gy and ∼7 Gy, the pixel values measured in portrait or landscape mode differed by 4.7%, 6.2% and 10.0%, respectively. A study of 200 film pieces revealed an excellent sheet-to-sheet uniformity. On a long time scale, the optical development of irradiated EBT film consisted of a slow but steady increase of absorbance which was not observed to cease during 4 months. Sensitometric curves of EBT films obtained under reference conditions (SSD = 95 cm, FS = 5 x 5 cm 2 , d = 5 cm) for 6, 10 and 25 MV photon beams did not show any energy dependence. The average separation between all curves was only 0.7%. The variation of the depth d (range 2-25 cm) in the phantom did not affect the dose response of EBT film. Also the influence of the radiation field size (range 3 x 3-40 x 40 cm 2 ) on the sensitometric curve was not significant. For EDR-2 films maximum differences between the calibration curves reached 7-8% for X6MV and X25MV. Radiochromic EBT film, in combination with a flatbed

  15. Spectral calibration of EBT3 and HD-V2 radiochromic film response at high dose using 20 MeV proton beams

    Science.gov (United States)

    Feng, Yiwei; Tiedje, Henry F.; Gagnon, Katherine; Fedosejevs, Robert

    2018-04-01

    Radiochromic film is used extensively in many medical, industrial, and scientific applications. In particular, the film is used in analysis of proton generation and in high intensity laser-plasma experiments where very high dose levels can be obtained. The present study reports calibration of the dose response of Gafchromic EBT3 and HD-V2 radiochromic films up to high exposure densities. A 2D scanning confocal densitometer system is employed to carry out accurate optical density measurements up to optical density 5 on the exposed films at the peak spectral absorption wavelengths. Various wavelengths from 400 to 740 nm are also scanned to extend the practical dose range of such films by measuring the response at wavelengths removed from the peak response wavelengths. Calibration curves for the optical density versus exposure dose are determined and can be used for quantitative evaluation of measured doses based on the measured optical densities. It was found that blue and UV wavelengths allowed the largest dynamic range though at some trade-off with overall accuracy.

  16. Dosimetric comparison between a planning system and Radiochromic-EBT2 films in surface brachytherapy treatments of high rate; Comparacion dosimetrica entre un sistema de planificacion y peliculas radiocromicas EBT2 en tratamientos de braquiterapia superficial de alta tasa

    Energy Technology Data Exchange (ETDEWEB)

    Gutierrez Ramos, S. M.; Carrasco Herrera, M. a.; Vicent, D.; Rodriguez, C.; Herrador, M.

    2013-07-01

    The objective of this work is to study a situation in which, the accuracy of the calculation of the planner may be limited, superficial brachytherapy treatment. It has relative to the dose obtained with the planner with that obtained with film radiochromic EBT2. (Author)

  17. Absorption spectra response of XRQA radiochromic film to x-ray radiation

    International Nuclear Information System (INIS)

    Alnawaf, Hani; Cheung, Tsang; Butson, Martin J.; Yu, Peter K.N.

    2010-01-01

    Gafchromic XRQA, radiochromic film is a high sensitivity auto developing x-ray analysis films designed and available for kilovoltage x-ray, dose and QA assessment applications. The film is designed for reflective analysis with a yellow transparent top filter and white opaque backing materials. This allows the film to be visually inspected for colour changes with a higher level of contrast than clear coated radiochromic films such as Gafchromic EBT version 1. The spectral absorption properties in the visible wavelengths have been investigated and results show two main peaks in absorption located at 636 nm and 585 nm. These peaks are located in the same position as EBT Gafchromic film highlighting a similar chemical monomer/polymer for radiation sensitivity. A much higher sensitivity however is found at kilovoltage energies with an average 1.55 OD units per 20 cGy irradiation variation measured at 636 nm using 150 kVp x-rays. This is compared to approximately 0.12 OD units per 20 cGy measured at 636 nm for EBT film at 6 MV x-ray energy. That is, the XRQA film is more than 10 times more sensitive than EBT1 film. The visual colour change is enhanced by the yellow polyester coating. However this does not affect the absorption spectra properties in the red region of analysis which is the main area for use using desktop scanners in reflection mode.

  18. SU-E-T-665: Radiochromic Film Quenching Effect Reduction for Proton Beam Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Aldelaijan, S; Alzorkany, F; Moftah, B; Alrumayan, F [King Faisal Specialist Hospital & Research Centre, Riyadh (Saudi Arabia); Seuntjens, J [McGill University, Montreal, QC (Canada); Lewis, D [RCF Consulting, LLC, Monroe, CT (United States); Devic, S [McGill University, Montreal, QC (Canada); Jewish General Hospital, Montreal, QC (Canada)

    2015-06-15

    Purpose: Depending on the useful dose range in which radiochromic films operate, number of different radiochromic film models have been designed. The impact of different film models on quenching effect for percent depth dose (PDD) measurements in proton beams has been investigated. Methods: Calibrated PTW Markus ionization chamber was used to measure PDD and beam output for 26.5 MeV protons produced by CS30 cyclotron. An aluminum cylinder was added in front of the beam exit serving as a radiation shutter. The measured signal was normalized to a monitor chamber reading and subsequently scaled by ratio of water-to-air stopping powers at given depth, while the effective depth of measurements was scaled by ratios of material-to-water physical densities and CSDA ranges. Output was measured in water at 2.1 mm reference-depth in the plateau upstream from the Bragg peak. Following the TRS-398 reference dosimetry protocol for proton beams, the output was calibrated in water. Three radiochromic film models (EBT, EBT3 and HD-V2) were calibrated within Lexan phantom positioned at the same water-equivalent depth. Thicknesses of films sensitive layers were 34 µm, 30 µm and 8 µm, respectively. Small film pieces (1 x 2 cm{sup 2}) were positioned within polyethylene phantom along the beam central axis with an angulation of 5° for PDD measurements. Results: While the output of the proton beam was found to be around 7 Gy/sec, the actual value of the output per monitor chamber reading (2.32 Gy/nC) was used for reference-dose irradiations during film calibration. Dose ratios at the Bragg peak relative to the reference-depth were 3.88, 2.52, 2.19, and 2.02 for the Markus chamber, HD-V2, EBT3, and EBT film models, respectively. Conclusion: Results at hand suggest that quenching effect is reduced when a radiochromic film model with smaller sensitive layer thickness is used for PDD measurements in proton beams. David Lewis is the owner of RCF Consulting, LLC.

  19. Evaluation of the Gafchromic® EBT2 film for the dosimetry of radiosurgical beams

    International Nuclear Information System (INIS)

    Lárraga-Gutiérrez, José M.; García-Hernández, Diana; García-Garduño, Olivia A.; Galván de la Cruz, Olga O.; Ballesteros-Zebadúa, Paola; Esparza-Moreno, Karina P.

    2012-01-01

    Purpose: Radiosurgery uses small fields and high-radiation doses to treat intra- and extracranial lesions in a single session. The lack of a lateral electronic equilibrium and the presence of high-dose gradients in these fields are challenges for adequate measurements. The availability of radiation detectors with the high spatial resolution required is restricted to only a few. Stereotactic diodes and EBT radiochromic films have been demonstrated to be good detectors for small-beam dosimetry. Because the stereotactic diode is the standard measurement for the dosimetry of radiosurgical beams, the goal of this work was to perform measurements with the radiochromic film Gafchromic ® EBT2 and compare its results with a stereotactic diode. Methods: Total scatter factors, tissue maximum, and off-axis ratios from a 6 MV small photon beams were measured using EBT2 radiochromic film in a water phantom. The film-measured data were evaluated by comparing it with the data measured with a stereotactic field diode (IBA-Dosimetry). Results: The film and diode measurements had excellent agreement. The differences between the detectors were less than or equal to 2.0% for the tissue maximum and the off-axis ratios. However, for the total scatter factors, there were significant differences, up to 4.9% (relative to the reference field), for field sizes less than 1.0 cm. Conclusions: This work found that the Gafchromic ® EBT2 film is adequate for small photon beam measurements, particularly for tissue maximum and off-axis ratios. However, careful attention must be taken when measuring output factors of small beams below 1.0 cm due to the film's energy dependence. The measurement differences may be attributable to the film's active layer composition because EBT2 incorporates higher Z elements (i.e., bromide and potassium), hence revealing a potential energy dependence for the dosimetry of small photon beams.

  20. Q A IMRT comparison specific patient by means of radiochromic films, radiographic films and ionization chambers arrangement

    International Nuclear Information System (INIS)

    Medina, L.; Venencia, D.; Garrigo, E.

    2014-08-01

    IMRT uses radiation beams of nonuniform intensity. Quality assurance (Q A) specific patient is mandatory in this treatment modality. The purpose of this study is to compare results of patient specific Q A IMRT dose distributions of the total plan and individual fields using different dosimetric systems. We used a photon beam 6 MV generated for linear accelerator PRIMUS, were used planning systems iPLAN and Konrad for IMRT inverse planning with modality Step and Shoot. For plans total dose distributions were measured with radiographic films EDR2 and Radiochromic Film EBT3. For individual fields the dose distributions were measured with radiographic films X-Omat-V, Radiochromic Film EBT3 and PTW 2D-Array. We used a scanner VIDAR Dosimetry Pro Red and software Rit v6.1 for analysis, was used Gamma index [Γ] for comparison of measured and calculated dose recording the number of pixels with Γ> 1. We analyzed 50 plan dose distributions total 50 individual fields. For the total plan the number of pixels with Γ>1 (3%-3m m) was 0.7%±1.2 [0.1%; 2.82%] for EBT3 y 1%±1.8 [0.2%; 3%] for EDR2. For individual fields (5%-3m m) was obtained 0.97%±1,7 [0%, 3%] for X-Omat-V, 0.84%±1.1[0.3%,3.1%] for EBT3 and 2.6%±1.9 [0.01%,6.8%] PTW 2D-Array. All three methods can be used. Radiochromic Films revealed the advantage and disadvantage of the cost. Both systems are slightly better film to PTW 2D-Array. (author)

  1. Investigation of EBT2 and EBT3 films for proton dosimetry in the 4-20 MeV energy range.

    Science.gov (United States)

    Reinhardt, S; Würl, M; Greubel, C; Humble, N; Wilkens, J J; Hillbrand, M; Mairani, A; Assmann, W; Parodi, K

    2015-03-01

    Radiochromic films such as Gafchromic EBT2 or EBT3 films are widely used for dose determination in radiation therapy because they offer a superior spatial resolution compared to any other digital dosimetric 2D detector array. The possibility to detect steep dose gradients is not only attractive for intensity-modulated radiation therapy with photons but also for intensity-modulated proton therapy. Their characteristic dose rate-independent response makes radiochromic films also attractive for dose determination in cell irradiation experiments using laser-driven ion accelerators, which are currently being investigated as future medical ion accelerators. However, when using these films in ion beams, the energy-dependent dose response in the vicinity of the Bragg peak has to be considered. In this work, the response of these films for low-energy protons is investigated. To allow for reproducible and background-free irradiation conditions, the films were exposed to mono-energetic protons from an electrostatic accelerator, in the 4-20 MeV energy range. For comparison, irradiation with clinical photons was also performed. It turned out that in general, EBT2 and EBT3 films show a comparable performance. For example, dose-response curves for photons and protons with energies as low as 11 MeV show almost no differences. However, corrections are required for proton energies below 11 MeV. Care has to be taken when correction factors are related to an average LET from depth-dose measurements, because only the dose-averaged LET yields similar results as obtained in mono-energetic measurements.

  2. Radiochromic film in the dosimetric verification of intensity modulated radiation therapy

    International Nuclear Information System (INIS)

    Zhou Yingjuan; Huang Shaomin; Deng Xiaowu

    2007-01-01

    Objective: Objective To investigate the dose-response behavior of a new type of radio- chromic film( GAFCHROMIC EBT) and explore the clinical application means and precision of dosage measurement, which can be applied for: (1) plan-specific dosimetric verification for intensity modulated radiation therapy, (2) to simplify the process of quality assurance using traditional radiographic film dosimetric system and (3) to establish a more reliable, more efficient dosimetric verification system for intensity modulated radiation therapy. Methods: (1) The step wedge calibration technique was used to calibrate EBT radiochromic film and EDR2 radiographic film. The dose characteristics, the measurement consistency and the quality assurance process between the two methods were compared. (2) The in-phantom dose-measurement based verification technique has been adopted. Respectively, EBT film and EDR2 film were used to measure the same dose plane of IMRT treatment plans. The results of the dose map, dose profiles and iso- dose curves were compared with those calculated by CORVUS treatment planning system to evaluate the function of EBT film for dosimetric verification for intensity modulated radiation therapy. Results: (1) Over the external beam dosimetric range of 0-500 cGy, EBT/VXR-16 and EDR2/VXR-16 film dosimetric system had the same measurement consistency with the measurement variability less then 0.70%. The mean measurement variability of these two systems was 0.37% and 0.68%, respectively. The former proved to be the superior modality at measurement consistency, reliability, and efficiency over dynamic clinical dose range , furthermore, its quality assurance showed less process than the latter. (2) The dosimetric verification of IMRT plane measured with EBT film was quite similar to that with EDR2 film which was processed under strict quality control. In a plane of the phantom, the maximal dose deviation off axis between EBT film measurement and the TPS calculation was

  3. Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density.

    Science.gov (United States)

    Marroquin, Elsa Y León; Herrera González, José A; Camacho López, Miguel A; Barajas, José E Villarreal; García-Garduño, Olivia A

    2016-09-08

    Radiochromic film has become an important tool to verify dose distributions for intensity-modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton's rings. Furthermore, the symmetrical design of EBT3 allows the user to eliminate side-orientation dependence. This film and the flatbed scanner, Epson Perfection V750, form a dosimetry system whose intrinsic characteristics were studied in this work. In addition, uncertainties associated with these intrinsic characteristics and the total uncertainty of the dosimetry system were determined. The analysis of the response of the radiochromic film (net optical density) and the fitting of the experimental data to a potential function yielded an uncertainty of 2.6%, 4.3%, and 4.1% for the red, green, and blue channels, respectively. In this work, the dosimetry system presents an uncertainty in resolving the dose of 1.8% for doses greater than 0.8 Gy and less than 6 Gy for red channel. The films irradiated between 0 and 120 Gy show differences in the response when scanned in portrait or landscape mode; less uncertainty was found when using the portrait mode. The response of the film depended on the position on the bed of the scanner, contributing an uncertainty of 2% for the red, 3% for the green, and 4.5% for the blue when placing the film around the center of the bed of scanner. Furthermore, the uniformity and reproducibility radiochromic film and reproducibility of the response of the scanner contribute less than 1% to the overall uncertainty in dose. Finally, the total dose uncertainty was 3.2%, 4.9%, and 5.2% for red, green, and blue channels, respectively. The above uncertainty values were obtained by mini-mizing the contribution to the total dose uncertainty

  4. Energy dependence of EBT-1 radiochromic film response for photon (10 kvp-15 MVp) and electron beams (6-18 MeV) readout by a flatbed scanner.

    Science.gov (United States)

    Richter, Christian; Pawelke, Jörg; Karsch, Leonhard; Woithe, Julia

    2009-12-01

    The aim of this article is to investigate the energy dependence of the radiochromic film type, Gafchromic EBT-1, when scanned with a flatbed scanner for film readout. Dose response curves were determined for 12 different beam qualities ranging from a 10 kVp x-ray beam to a 15 MVp x-ray beam and include also two high energy electron beam qualities (6 and 18 MeV). The dose responses measured as net optical density (netOD) for the different beam qualities were normalized to the response of a reference beam quality (6 MVp). A strong systematic energy dependence of the film response was found. The lower the effective beam energy, the less sensitive the EBT-1 films get. The maximum decrease in dose for the same film response between the 25 kVp and 6 MVp beam qualities was 44%. Additionally, a difference in energy dependence for different doses was discovered, meaning that higher doses show a smaller dependency on energy than lower doses. The maximum decrease in the normalized netOD was found to be 25% for a dose of 0.5 Gy relative to the normalized netOD for 10 Gy. Moreover, a scaling procedure is introduced, allowing the correction of the energy dependence for the investigated beam qualities and also for comparable x-ray beam qualities within the energy range studied. A strong energy dependence for EBT-1 radiochromic films was found. The films were readout with a flatbed scanner. If the effective beam energy is known, the energy dependence can be corrected with the introduced scaling procedure. Further investigation of the influence of the spectral band of the readout device on energy dependence is needed to understand the reason for the different energy dependences found in this and previous works.

  5. EXPERIMENTAL EVALUATION OF DOSIMETRIC CHARACTERIZATION OF GAFCHROMIC EBT3 AND EBT-XD FILMS FOR CLINICAL CARBON ION BEAMS.

    Science.gov (United States)

    Yonai, Shunsuke; Arai, Chinatsu; Shimoyama, Kaoru; Fournier-Bidoz, Nathalie

    2018-02-03

    Radiochromic film is a very useful tool for 2D dosimetric measurements in radiotherapy because it is self-developing and has very high-spatial resolution. However, considerable care has to be taken in ion beam radiotherapy owing to the quenching effect of high-linear energy transfer (LET) radiation. In this study, the dose responses of GAFchromic EBT3 and EBT-XD films were experimentally investigated using the clinical carbon ion beam at the Heavy Ion Medical Accelerator in Chiba. Results showed that the relations between absorbed dose and net optical density could be expressed well using an equation proposed by Reinhardt (2015). The quenching effect was evaluated by determining their relative efficiencies for photon irradiation as a function of LET. A correction equation derived in this study allowed the absorbed dose to be determined in the small irradiation field used for carbon ion radiotherapy eye treatments. This study contributes to establishing an absolute dosimetry procedure for heavy ion beams using radiochromic film. © The Author(s) 2018. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  6. SU-E-T-485: In Vivo Dosimetry with EBT3 Radiochromic Films for TBI Treatments

    Energy Technology Data Exchange (ETDEWEB)

    Lozares, S; Gracia, M; Olasolo, J; Gallardo, N; Fuentemilla, N; Pellejero, S; Miquelez, S; Maneru, F; Martin, M; Bragado, L; Rubio, A [Complejo Hospitalario de Navarra, Pamplona, Navarra (Spain)

    2015-06-15

    Purpose: Total body irradiation (TBI) is a technique that requires special equipment to control “in vivo” the dose to the patient because it is a complex technique performed in extraordinary conditions. There are several devices to perform this task (diodes, TLDs, ionization chambers, MOSFET). In this paper we study the possibility of performing these measurements with radiochromic films EBT3 properly calibrated. This method has been compared to the PTW diodes system for TBI. Methods: Once made the TC to the patients, we measured different thicknesses of the relevant areas of the body (head, neck, chest with or without arms, umbilicus area, knees and ankles); for each of these thicknesses we measured dose rate (cGy / UM) in RW3 phantom, in TBI conditions, with ionization chamber in the center; in turn, the input diode and the output of each configuration is placed to assign dose to each set of diodes. Movie calibration is performed according to manufacturer’s recommendations but TBI conditions. The dose at the center of each thickness compared to a linear interpolation of the dose at the entrance and exit, resulting in an adequate approximation. Finally in each session for each patient put a piece of film (2×2 cm2) at the entrance and another at the exit in each area, obtaining these readings and interpolating the estimated center dose, as with the diodes. Results: These results show a greater homogeneity in the distribution for use with film and validate the use of the same for this task and, if necessary, to avoid purchasing diode group if they have not. Conclusion: By using radiochromic films for this technique gives us a proper calculation of the dose received by the patient in the absence of other methods, or gives us a second additional track that already used normally.

  7. SU-E-T-485: In Vivo Dosimetry with EBT3 Radiochromic Films for TBI Treatments

    International Nuclear Information System (INIS)

    Lozares, S; Gracia, M; Olasolo, J; Gallardo, N; Fuentemilla, N; Pellejero, S; Miquelez, S; Maneru, F; Martin, M; Bragado, L; Rubio, A

    2015-01-01

    Purpose: Total body irradiation (TBI) is a technique that requires special equipment to control “in vivo” the dose to the patient because it is a complex technique performed in extraordinary conditions. There are several devices to perform this task (diodes, TLDs, ionization chambers, MOSFET). In this paper we study the possibility of performing these measurements with radiochromic films EBT3 properly calibrated. This method has been compared to the PTW diodes system for TBI. Methods: Once made the TC to the patients, we measured different thicknesses of the relevant areas of the body (head, neck, chest with or without arms, umbilicus area, knees and ankles); for each of these thicknesses we measured dose rate (cGy / UM) in RW3 phantom, in TBI conditions, with ionization chamber in the center; in turn, the input diode and the output of each configuration is placed to assign dose to each set of diodes. Movie calibration is performed according to manufacturer’s recommendations but TBI conditions. The dose at the center of each thickness compared to a linear interpolation of the dose at the entrance and exit, resulting in an adequate approximation. Finally in each session for each patient put a piece of film (2×2 cm2) at the entrance and another at the exit in each area, obtaining these readings and interpolating the estimated center dose, as with the diodes. Results: These results show a greater homogeneity in the distribution for use with film and validate the use of the same for this task and, if necessary, to avoid purchasing diode group if they have not. Conclusion: By using radiochromic films for this technique gives us a proper calculation of the dose received by the patient in the absence of other methods, or gives us a second additional track that already used normally

  8. Energy dependence of EBT-1 radiochromic film response for photon (10 kVp-15 MVp) and electron beams (6-18 MeV) readout by a flatbed scanner

    International Nuclear Information System (INIS)

    Richter, Christian; Pawelke, Joerg; Karsch, Leonhard; Woithe, Julia

    2009-01-01

    Purpose: The aim of this article is to investigate the energy dependence of the radiochromic film type, Gafchromic EBT-1, when scanned with a flatbed scanner for film readout. Methods: Dose response curves were determined for 12 different beam qualities ranging from a 10 kVp x-ray beam to a 15 MVp x-ray beam and include also two high energy electron beam qualities (6 and 18 MeV). The dose responses measured as net optical density (netOD) for the different beam qualities were normalized to the response of a reference beam quality (6 MVp). Results: A strong systematic energy dependence of the film response was found. The lower the effective beam energy, the less sensitive the EBT-1 films get. The maximum decrease in dose for the same film response between the 25 kVp and 6 MVp beam qualities was 44%. Additionally, a difference in energy dependence for different doses was discovered, meaning that higher doses show a smaller dependency on energy than lower doses. The maximum decrease in the normalized netOD was found to be 25% for a dose of 0.5 Gy relative to the normalized netOD for 10 Gy. Moreover, a scaling procedure is introduced, allowing the correction of the energy dependence for the investigated beam qualities and also for comparable x-ray beam qualities within the energy range studied. Conclusions: A strong energy dependence for EBT-1 radiochromic films was found. The films were readout with a flatbed scanner. If the effective beam energy is known, the energy dependence can be corrected with the introduced scaling procedure. Further investigation of the influence of the spectral band of the readout device on energy dependence is needed to understand the reason for the different energy dependences found in this and previous works.

  9. EBT-XD Radiochromic Film Sensitivity Calibrations Using Proton Beams from a Pelletron Accelerator

    Science.gov (United States)

    Stockler, Barak; Grun, Alexander; Brown, Gunnar; Klein, Matthew; Wood, Jacob; Cooper, Anthony; Ward, Ryan; Freeman, Charlie; Padalino, Stephen; Regan, S. P.; Sangster, T. C.

    2017-10-01

    Radiochromic film (RCF) is a transparent detector film that permanently changes color following exposure to ionizing radiation. RCF is used frequently in medical applications, but also has been used in a variety of high energy density physics diagnostics. RCF is convenient to use because it requires no chemical processing and can be scanned using commercially available document scanners. In this study, the sensitivity of Gafchromic™ EBT-XD RCF to protons and x-rays was measured. Proton beams produced by the SUNY Geneseo Pelletron accelerator were directed into an evacuated target chamber where they scattered off a thin gold foil. The scattered protons were incident on a sample of RCF which subtended a range of angles around the scattering center. A new analysis method, which relies on the variation in scattered proton fluence as a function of scattering angle in accordance with the Rutherford scattering law, is currently being developed to speed up the proton calibrations. Samples of RCF were also exposed to x-ray radiation using an X-RAD 160 x-ray irradiator, allowing the sensitivity of RCF to X-rays to be measured. This work was funded in part by a Grant from the DOE through the Laboratory for Laser Energetics as well as the NSF.

  10. Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density

    Science.gov (United States)

    Marroquin, Elsa Y. León; Herrera González, José A.; Camacho López, Miguel A.; Barajas, José E. Villarreal

    2016-01-01

    Radiochromic film has become an important tool to verify dose distributions for intensity‐modulated radiotherapy (IMRT) and quality assurance (QA) procedures. A new radiochromic film model, EBT3, has recently become available, whose composition and thickness of the sensitive layer are the same as those of previous EBT2 films. However, a matte polyester layer was added to EBT3 to prevent the formation of Newton's rings. Furthermore, the symmetrical design of EBT3 allows the user to eliminate side‐orientation dependence. This film and the flatbed scanner, Epson Perfection V750, form a dosimetry system whose intrinsic characteristics were studied in this work. In addition, uncertainties associated with these intrinsic characteristics and the total uncertainty of the dosimetry system were determined. The analysis of the response of the radiochromic film (net optical density) and the fitting of the experimental data to a potential function yielded an uncertainty of 2.6%, 4.3%, and 4.1% for the red, green, and blue channels, respectively. In this work, the dosimetry system presents an uncertainty in resolving the dose of 1.8% for doses greater than 0.8 Gy and less than 6 Gy for red channel. The films irradiated between 0 and 120 Gy show differences in the response when scanned in portrait or landscape mode; less uncertainty was found when using the portrait mode. The response of the film depended on the position on the bed of the scanner, contributing an uncertainty of 2% for the red, 3% for the green, and 4.5% for the blue when placing the film around the center of the bed of scanner. Furthermore, the uniformity and reproducibility radiochromic film and reproducibility of the response of the scanner contribute less than 1% to the overall uncertainty in dose. Finally, the total dose uncertainty was 3.2%, 4.9%, and 5.2% for red, green, and blue channels, respectively. The above uncertainty values were obtained by minimizing the contribution to the total dose

  11. Evaluation of gafchromic EBT film for intensity modulated radiation therapy dose distribution verification

    International Nuclear Information System (INIS)

    Sankar, A.; Gopalkrishna Kurup, P.G.; Murali, V.; Ayyangar, Komanduri M.; Mothilal Nehru, R.; Velmurugan, J.

    2006-01-01

    This work was undertaken with the intention of investigating the possibility of clinical use of commercially available self-developing radiochromic film - Gafchromic EBT film - for IMRT dose verification. The dose response curves were generated for the films using VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak EDR2 films. It was found that the EBT film has a linear response between the dose ranges of 0 and 600 cGy. The dose-related characteristics of the EBT film, like post-irradiation color growth with time, film uniformity and effect of scanning orientation, were studied. There is up to 8.6% increase in the color density between 2 and 40 h after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative difference between calculated and measured dose distributions was analyzed using Gamma index with the tolerance of 3% dose difference and 3 mm distance agreement. EDR2 films showed good and consistent results with the calculated dose distribution, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large field IMRT verification. For IMRT of smaller field size (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films. (author)

  12. A feasibility study using radiochromic films for fast neutron 2D passive dosimetry

    International Nuclear Information System (INIS)

    Brady, Samuel L; Fallin, Brent; Gunasingha, Rathnayaka; Yoshizumi, Terry T; Howell, Calvin R; Crowell, Alexander S; Tonchev, Anton P; Dewhirst, Mark W

    2010-01-01

    The objective of this paper is threefold: (1) to establish sensitivity of XRQA and EBT radiochromic films to fast neutron exposure; (2) to develop a film response to radiation dose calibration curve and (3) to investigate a two-dimensional (2D) film dosimetry technique for use in establishing an experimental setup for a radiobiological irradiation of mice and to assess the dose to the mice in this setup. The films were exposed to a 10 MeV neutron beam via the 2 H(d,n) 3 He reaction. The XRQA film response was a factor of 1.39 greater than EBT film response to the 10 MeV neutron beam when exposed to a neutron dose of 165 cGy. A film response-to-soft tissue dose calibration function was established over a range of 0-10 Gy and had a goodness of fit of 0.9926 with the calibration data. The 2D film dosimetry technique estimated the neutron dose to the mice by measuring the dose using a mouse phantom and by placing a piece of film on the exterior of the experimental mouse setup. The film results were benchmarked using Monte Carlo and aluminum (Al) foil activation measurements. The radiochromic film, Monte Carlo and Al foil dose measurements were strongly correlated, and the film within the mouse phantom agreed to better than 7% of the externally mounted films. These results demonstrated the potential application of radiochromic films for passive 2D neutron dosimetry.

  13. Evaluation of the Gafchromic{sup Registered-Sign} EBT2 film for the dosimetry of radiosurgical beams

    Energy Technology Data Exchange (ETDEWEB)

    Larraga-Gutierrez, Jose M. [Laboratorio de Fisica Medica, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Mexico D.F. 14269 (Mexico) and Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, C.P. 14269, Mexico D.F. 14269 (Mexico); Garcia-Hernandez, Diana [Laboratorio de Fisica Medica, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Mexico D.F. 14269 (Mexico); Garcia-Garduno, Olivia A. [Laboratorio de Fisica Medica, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Mexico D.F. 14269 (Mexico) and Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Unidad Legaria, Instituto Politecnico Nacional, Legaria 694, Mexico D.F. 11500 (Mexico); Galvan de la Cruz, Olga O. [Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Mexico D.F. 14269 (Mexico); Ballesteros-Zebadua, Paola [Laboratorio de Fisica Medica, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Mexico D.F. 14269 (Mexico) and Unidad de Radioneurocirugia, Instituto Nacional de Neurologia y Neurocirugia, Insurgentes Sur 3877, Mexico D.F. 14269 (Mexico); Esparza-Moreno, Karina P. [Facultad de Medicina, Universidad Autonoma del Estado de Mexico, Paseo Tollocan, Toluca, Estado De Mexico 50180 (Mexico)

    2012-10-15

    Purpose: Radiosurgery uses small fields and high-radiation doses to treat intra- and extracranial lesions in a single session. The lack of a lateral electronic equilibrium and the presence of high-dose gradients in these fields are challenges for adequate measurements. The availability of radiation detectors with the high spatial resolution required is restricted to only a few. Stereotactic diodes and EBT radiochromic films have been demonstrated to be good detectors for small-beam dosimetry. Because the stereotactic diode is the standard measurement for the dosimetry of radiosurgical beams, the goal of this work was to perform measurements with the radiochromic film Gafchromic{sup Registered-Sign} EBT2 and compare its results with a stereotactic diode. Methods: Total scatter factors, tissue maximum, and off-axis ratios from a 6 MV small photon beams were measured using EBT2 radiochromic film in a water phantom. The film-measured data were evaluated by comparing it with the data measured with a stereotactic field diode (IBA-Dosimetry). Results: The film and diode measurements had excellent agreement. The differences between the detectors were less than or equal to 2.0% for the tissue maximum and the off-axis ratios. However, for the total scatter factors, there were significant differences, up to 4.9% (relative to the reference field), for field sizes less than 1.0 cm. Conclusions: This work found that the Gafchromic{sup Registered-Sign} EBT2 film is adequate for small photon beam measurements, particularly for tissue maximum and off-axis ratios. However, careful attention must be taken when measuring output factors of small beams below 1.0 cm due to the film's energy dependence. The measurement differences may be attributable to the film's active layer composition because EBT2 incorporates higher Z elements (i.e., bromide and potassium), hence revealing a potential energy dependence for the dosimetry of small photon beams.

  14. Local heterogeneities in early batches of EBT2 film: a suggested solution.

    Science.gov (United States)

    Kairn, T; Aland, T; Kenny, J

    2010-08-07

    To enhance the utility of radiochromic films for high-resolution dosimetry of small and modulated radiotherapy fields, we propose a means to negate the effects of heterogeneities in EBT2 (and other) films. The results of using our simple procedure for evaluating radiation dose in EBT2 film are compared with the results of using the manufacturer's recommended procedure as well as a procedure previously established for evaluating dose in older EBT film. It is shown that Newton's ring-like scanning artefacts can be avoided through the use of a plastic frame, to elevate the film above the scanner's surface. The effects of film heterogeneity can be minimized by evaluating net optical density, pixelwise, as the logarithm of the ratio of the red-channel pixel value in each pixel of each irradiated film to the red-channel pixel value in the same pixel in the same film prior to irradiation. The application of a blue-channel correction was found to result in increased noise. It is recommended that, when using EBT2 film for radiotherapy quality assurance, the films should be scanned before and after irradiation and analysed using the method proposed herein, without the use of the blue-channel correction, in order to produce dose images with minimal film heterogeneity effects.

  15. Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

    International Nuclear Information System (INIS)

    Lewis, David F.; Chan, Maria F.

    2016-01-01

    Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners

  16. Technical Note: On GAFChromic EBT-XD film and the lateral response artifact.

    Science.gov (United States)

    Lewis, David F; Chan, Maria F

    2016-02-01

    The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners. Owing its lower sensitivity, EBT-XD film

  17. Technical Note: On GAFChromic EBT-XD film and the lateral response artifact

    Energy Technology Data Exchange (ETDEWEB)

    Lewis, David F., E-mail: rcfilmconsulting@gmail.com [RC Film Consulting LLC, 54 Benedict Road, Monroe, Connecticut 06468 (United States); Chan, Maria F. [Memorial Sloan-Kettering Cancer Center, 136 Mountain View Boulevard, Basking Ridge, New Jersey 07920 (United States)

    2016-02-15

    Purpose: The new radiochromic film, GAFChromic EBT-XD, contains the same active material, lithium-10,12-pentacosadiynoate, as GAFChromic EBT3, but the crystalline form is different. This work investigates the effect of this change on the well-known lateral response artifact when EBT-XD film is digitized on a flatbed scanner. Methods: The dose response of a single production lot of EBT-XD was characterized by scanning an unexposed film plus a set of films exposed to doses between 2.5 and 50 Gy using 6 MV photons. To characterize the lateral response artifact, the authors used the unexposed film plus a subset of samples exposed to doses between 20 and 50 Gy. Digital images of these films were acquired at seven discrete lateral locations perpendicular to the scan direction on three Epson 10000XL scanners. Using measurements at the discrete lateral positions, the scanner responses were determined as a function of the lateral position of the film. From the data for each scanner, a set of coefficients were derived whereby measured response values could be corrected to remove the effects of the lateral response artifact. The EBT-XD data were analyzed as in their previous work and compared to results reported for EBT3 in that paper. Results: For films scanned in the same orientation and having equal responses, the authors found that the lateral response artifact for EBT-XD and EBT3 films was remarkably similar. For both films, the artifact increases with increased net response. However, as EBT-XD is less sensitive than EBT3, a greater exposure dose is required to reach the same net response. On this basis, the lower sensitivity of EBT-XD relative to EBT3 results in less net response change for equal exposure and a reduction in the impact of the lateral response artifact. Conclusions: The shape of the crystalline active component in EBT-XD and EBT3 does not affect the fundamental existence of the lateral response artifact when the films are digitized on flatbed scanners

  18. Characterization of beta radiation fields using radiochromic films

    International Nuclear Information System (INIS)

    Benavente, Jhonny A.; Silva, Teogenes A. da

    2011-01-01

    The objective of this work was to study the response of radiochromic films for beta radiation fields in terms of absorbed dose. The reliability of the EBT model Gafchromic radiochromic film was studied. A 9800 XL model Microtek, transmission scanner, a 369 model X-Rite optical densitometer and a Mini 1240 Shimadzu UV spectrophotometer were used for measurement comparisons. Calibration of the three systems was done with irradiated samples of radiochromic films with 0.1; 0.3; 0.5; 0.8; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.5 e 5.0 Gy in beta radiation field from a Sr-90/Y-90 source. Calibration was performed by establishing a correlation between the absorbed dose values and the corresponding radiochromic responses. Results showed significant differences in the absorbed dose values obtained with the three methods. Absorbed dose values showed errors from 0.6 to 4.4%, 0.3 to 31.8% and 0.2 to 47.3% for the Microtek scanner, the X-Rite Densitometer and the Shimadzu spectrophotometer, respectively. Due to the easy acquisition and use for absorbed dose measurements, the densitometer and the spectrophotometer showed to be suitable techniques to evaluate radiation dose in relatively homogeneous fields. In the case of inhomogeneous fields or for a two dimension mapping of radiation fields to identify anisotropies, the scanner technique is the most recommended. (author)

  19. Solar ultraviolet radiation response of EBT2 Gafchromic, radiochromic film

    International Nuclear Information System (INIS)

    Butson, Ethan T; Yu, Peter K N; Butson, Martin J

    2013-01-01

    Measurement of solar ultraviolet (UV) radiation is an important aspect of dosimetry for the improved knowledge of UV exposure and its associated health related issues. EBT2 Gafchromic film has been designed by its manufacturers as an improved tool for ionizing radiation dosimetry. The film is stated as exhibiting a significant reduction in UV response. However, results have shown that when exposed to UV from the ‘bottom side’ i.e. from the thick laminate side, the film exhibits a sensitivity to solar UV radiation which is both measurable and accurate for UV dosimetry. Films were irradiated in this position to known solar UV exposures and results are quantified showing a reproducibility of measurement to within ±7% (1 SD) when compared to calibrated UV meters. With an exposure of 20 J cm −2 broad spectrum solar UV, the films net OD change was found to be 0.248 OD ± 0.021 OD when analysing the results using the red channel region of an Epson V700 desktop scanner. This was compared to 0.0294 OD ± 0.0053 OD change with exposure to the same UV exposure from the top side. This means that solar UV dosimetry can be performed using EBT2 Gafchromic film utilizing the underside of the film for dosimetry. The main advantages of this film type for measurement of UV exposure is the visible colour change and thus easy analysis using a desktop scanner as well as its uniformity in response and its robust physical strength for use in outside exposure situations. (note)

  20. SU-F-T-573: Evaluation of EBT-XD Radiochromic Films for Verification of SBRT and SRS Treatment Delivery

    Energy Technology Data Exchange (ETDEWEB)

    Aubry, J; Zerouali, K [Centre hospitalier de l’Universite de Montreal, Montreal, Quebec (Canada)

    2016-06-15

    Purpose: To evaluate the accuracy and precision of radiochromic films EBT-XD for quality control of stereotaxic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) plan delivery. Methods: A film cut in 3×10 cm2 strips was irradiated from 0 to 20 Gy in increments of 1 to 1.5 Gy using a 15 MeV electron beam. Triple-channel film calibration was completed 24 hours later by scanning the film strips on an Epson 10000XL scanner using a well-defined protocol. Several dose measurements of increasing complexity were subsequently performed with Varian iX accelerators. Pieces of films were first irradiated in a solid water phantom with 6 MV photon beams and a static gantry to doses spanning the calibration range, either in a single field or multiple fields setup. High dose (>15 Gy per fraction) IMRT plans were then measured. Finally films were irradiated with volumetric modulated arc therapy (VMAT) plans of lung and spinal lesions with prescribed doses per fraction between 8 and 20 Gy. The dose measured with the films was compared to the calculated dose from the Eclipse planning system using the Anisotropic Analytical Algorithm (AAA). Results: 77 dose measurements were compared to either ion chamber measurements or dose calculations (reference). The average dose difference between film measurements and reference was 0.7 % and the standard deviation was 1.3%. The maximum and minimum dose differences were +3.5% and −2% in the 4 Gy to 20 Gy range. Measured dose profiles of lung and vertebra treatment plans agreed very well with the calculations. Conclusion: EBT-XD films are a useful dosimeter for quality control of SBRT and SRS plan delivery. The measurement of a full 2D dose plane with high spatial resolution and acceptable dose accuracy make it an advantageous choice compared to other detectors such as ion chambers or diodes.

  1. SU-E-T-135: Investigation of Commercial-Grade Flatbed Scanners and a Medical- Grade Scanner for Radiochromic EBT Film Dosimetry.

    Science.gov (United States)

    Syh, J; Patel, B; Syh, J; Wu, H; Rosen, L; Durci, M; Katz, S; Sibata, C

    2012-06-01

    To evaluate the characteristics of commercial-grade flatbed scanners and medical-grade scanners for radiochromic EBT film dosimetry. Performance aspects of a Vidar Dosimetry Pro Advantage (Red), Epson 750 Pro, Microtek ArtixScan 1800f, and Microtek ScanMaker 8700 scanner for EBT2 Gafchromic film were evaluated in the categories of repeatability, maximum distinguishable optical density (OD) differentiation, OD variance, and dose curve characteristics. OD step film by Stouffer Industries containing 31 steps ranging from 0.05 to 3.62 OD was used. EBT films were irradiated with dose ranging from 20 to 600 cGy in 6×6 cm 2 field sizes and analyzed 24 hours later using RIT113 and Tomotherapy Film Analyzer software. Scans were performed in transmissive mode, landscape orientation, 16-bit image. The mean and standard deviation Analog to Digital (A/D) scanner value was measured by selecting a 3×3 mm 2 uniform area in the central region of each OD step from a total of 20 scans performed over several weeks. Repeatability was determined from the variance of OD step 0.38. Maximum distinguishable OD was defined as the last OD step whose range of A/D values does not overlap with its neighboring step. Repeatability uncertainty ranged from 0.1% for Vidar to 4% for Epson. Average standard deviation of OD steps ranged from 0.21% for Vidar to 6.4% for ArtixScan 1800f. Maximum distinguishable optical density ranged from 3.38 for Vidar to 1.32 for ScanMaker 8700. A/D range of each OD step corresponds to a dose range. Dose ranges of OD steps varied from 1% for Vidar to 20% for ScanMaker 8700. The Vidar exhibited a dose curve that utilized a broader range of OD values than the other scanners. Vidar exhibited higher maximum distinguishable OD, smaller variance in repeatability, smaller A/D value deviation per OD step, and a shallower dose curve with respect to OD. © 2012 American Association of Physicists in Medicine.

  2. Quality assurance of alpha-particle dosimetry using peeled-off Gafchromic EBTfilm

    Science.gov (United States)

    Ng, C. Y. P.; Chun, S. L.; Yu, K. N.

    2016-08-01

    A novel alpha-particle dosimetry technique using Gafchromic EBT3 film has recently been proposed for calibrating the activity of alpha-emitting radiopharmaceuticals. In the present paper, we outlined four measures which could further help assure the quality of the method. First, we suggested an alternative method in fabricating the peeled-off EBT3 film. Films with a chosen size were cut from the original films and all the edges were sealed with silicone. These were immersed into deionized water for 19 d and the polyester covers of the EBT3 films could then be easily peeled off. The active layers in these peeled-off EBT3 films remained intact, and these films could be prepared reproducibly with ease. Second, we proposed a check on the integrity of the peeled-off film by comparing the responses of the pristine and peeled-off EBT3 films to the same X-ray irradiation. Third, we highlighted the importance of scanning directions of the films. The ;landscape; and ;portrait; scanning directions were defined as the scanning directions perpendicular and parallel to the long edge of the original EBT3 films, respectively. Our results showed that the responses were different for different scanning directions. As such, the same scanning direction should be used every time. Finally, we cautioned the need to confirm the uniformity of the alpha-particle source used for calibration. Radiochromic films are well known for their capability of providing two-dimensional dosimetric information. As such, EBT3 films could also be conveniently used to check the uniformity of the alpha-particle source.

  3. Q A IMRT comparison specific patient by means of radiochromic films, radiographic films and ionization chambers arrangement; Comparacion de QA IMRT paciente especifico mediante films radiocromicos, films radiograficos y arreglo de camaras de ionizacion

    Energy Technology Data Exchange (ETDEWEB)

    Medina, L.; Venencia, D.; Garrigo, E., E-mail: fisicamedina11@gmail.com [Instituto Privado de Radioterapia, Obispo Oro 423, X5000BFI Cordoba (Argentina)

    2014-08-15

    IMRT uses radiation beams of nonuniform intensity. Quality assurance (Q A) specific patient is mandatory in this treatment modality. The purpose of this study is to compare results of patient specific Q A IMRT dose distributions of the total plan and individual fields using different dosimetric systems. We used a photon beam 6 MV generated for linear accelerator PRIMUS, were used planning systems iPLAN and Konrad for IMRT inverse planning with modality Step and Shoot. For plans total dose distributions were measured with radiographic films EDR2 and Radiochromic Film EBT3. For individual fields the dose distributions were measured with radiographic films X-Omat-V, Radiochromic Film EBT3 and PTW 2D-Array. We used a scanner VIDAR Dosimetry Pro Red and software Rit v6.1 for analysis, was used Gamma index [Γ] for comparison of measured and calculated dose recording the number of pixels with Γ> 1. We analyzed 50 plan dose distributions total 50 individual fields. For the total plan the number of pixels with Γ>1 (3%-3m m) was 0.7%±1.2 [0.1%; 2.82%] for EBT3 y 1%±1.8 [0.2%; 3%] for EDR2. For individual fields (5%-3m m) was obtained 0.97%±1,7 [0%, 3%] for X-Omat-V, 0.84%±1.1[0.3%,3.1%] for EBT3 and 2.6%±1.9 [0.01%,6.8%] PTW 2D-Array. All three methods can be used. Radiochromic Films revealed the advantage and disadvantage of the cost. Both systems are slightly better film to PTW 2D-Array. (author)

  4. Radiochromic film for individual patient QA in extracranial stereotactic lung radiotherapy

    International Nuclear Information System (INIS)

    Kron, T.; Clements, N.; Aarons, Y.; Dunn, L.; Chesson, B.; Miller, J.; Roozen, K.; Ball, D.

    2011-01-01

    Introduction: Modern radiotherapy is characterised by increasingly complex radiation delivery such as Intensity Modulated Radiation Therapy (IMRT) or extracranial stereotactic radiotherapy (ESR). It has become common practice to verify the delivery for each patient in IMRT, however, no such methods have been defined to date for ESR. It was the aim of the present work to develop a method to verify the dose distribution for ESR in a moving phantom using radiochromic film. Methods: Radiochromic film (ISP EBT2) was used in a cylindrical film cassette fitted into a QUASAR phantom (Modus Medical). The cassette can be moved forwards and backwards with motion patterns that can mimic the breathing of individual patients. The radiotherapy treatment plans of four patients were re-planned for the phantom. Between 8 and 10 radiation fields of 6 MV photons from a Varian Trilogy linear accelerator were used to deliver target doses between 18 and 26 Gy per fraction. In order to allow for measurements with EBT2 film all monitor units were divided by 3 or 4 resulting in maximum doses not exceeding 10 Gy. The film was evaluated using a transmission scan on an Epson Perfection V700 scanner with 50 dpi spatial resolution (3 colors, 48bit). A calibration curve was used to convert either the red or green component of the scan to dose. Results: Qualitative film analysis found no discernible discrepancies from planned isodose distributions in the stationary images. The dose distribution in the moving phantom was used successfully to confirm the appropriateness of the ITV construction in the planning process. Conclusions: We have developed a QA procedure that accounts for breathing patterns of individual patients in ESR for lung cancer. Radiochromic film was found to be easy to adapt to this process.

  5. Evaluation of radiochromic films EBT3 for in-vivo dosimetry in radiotherapy treatments with photons

    International Nuclear Information System (INIS)

    Galvan de la C, O. O.; Rivera M, T.; Garcia G, O. A.; Larraga G, J. M.

    2015-10-01

    Full text: In-vivo dosimetry is a challenge in radiotherapy due to the measures are carried out in reference conditions outside; there is no balance of charged particle and beam consists of photons own and contamination electrons. Detectors that are useful for such measures should be sufficiently small and thin so they do not disturb the beam and do not alter the dose on target. In this paper the radiochromic films Gafchromic model EBT3 are evaluated as potential detectors for in-vivo dosimetry; measurements were carried out in solid water phantom on the surface, with films of dimensions 3 x 3 cm 2 . Irradiations were performed with a linear accelerator Novalis of 6 MV. Comparison between dose values found with a diode type Sfd detector (IBA dosimetry, Germany) and a diamond detector CVD (PTW-Freiburg) for 2 different sized of field (5 x 5 cm 2 and 10 x 10 cm 2 ) on the surface of a water phantom scanning was realized. The total spreading factor (Tsf) measured on surface was of 0.831 ± 4.6%; which is greater 12.9% than Tsf measured at a depth of maximum dose. This difference may be due to the contribution of scattered electrons to the beam exit. The measures must be corroborated with Monte Carlo simulations, which they will be validated on surface by the Abdel-Rahman [et al.] method; this mechanism will determine if the films are useful detectors for in-vivo dosimetry clinically. (Author)

  6. Quality assurance of alpha-particle dosimetry using peeled-off Gafchromic EBTfilm

    International Nuclear Information System (INIS)

    Ng, C.Y.P.; Chun, S.L.; Yu, K.N.

    2016-01-01

    A novel alpha-particle dosimetry technique using Gafchromic EBT3 film has recently been proposed for calibrating the activity of alpha-emitting radiopharmaceuticals. In the present paper, we outlined four measures which could further help assure the quality of the method. First, we suggested an alternative method in fabricating the peeled-off EBT3 film. Films with a chosen size were cut from the original films and all the edges were sealed with silicone. These were immersed into deionized water for 19 d and the polyester covers of the EBT3 films could then be easily peeled off. The active layers in these peeled-off EBT3 films remained intact, and these films could be prepared reproducibly with ease. Second, we proposed a check on the integrity of the peeled-off film by comparing the responses of the pristine and peeled-off EBT3 films to the same X-ray irradiation. Third, we highlighted the importance of scanning directions of the films. The “landscape” and “portrait” scanning directions were defined as the scanning directions perpendicular and parallel to the long edge of the original EBT3 films, respectively. Our results showed that the responses were different for different scanning directions. As such, the same scanning direction should be used every time. Finally, we cautioned the need to confirm the uniformity of the alpha-particle source used for calibration. Radiochromic films are well known for their capability of providing two-dimensional dosimetric information. As such, EBT3 films could also be conveniently used to check the uniformity of the alpha-particle source. - Highlights: • Proposed method to fabricate peeled-off EBT3 films for alpha dosimetry. • Proposed integrity check of peeled-off EBT3 films using X-ray irradiation. • Highlighted importance of scanning directions of EBT3 films. • Cautioned the need for uniformity check on alpha-particle source.

  7. Development and implementation of own software for dosimetry multichannel film; Desarrollo e implementacion de un software propio para la dosimetria multicanal con pelicula radiocromica EBT2

    Energy Technology Data Exchange (ETDEWEB)

    Jimenez Feltstrom, D.; Reyes Garcia, R.; Luis Simon, F. J.; Carrasco Herrera, M.; Sanchez Carmona, G.; Herrador Cordoba, M.

    2013-07-01

    The objective of this work is to develop its own software for multichannel film dosimetry Radiochromic EBT2. Compare the results obtained with its use in multichannel and single-channel dosimetry. Check that the multi-channel dosimetry eliminates much of the artifacts caused by dirt, fingerprints, scratches, etc. Radiochromic in film and scanner devices. (Author)

  8. The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers.

    Science.gov (United States)

    Schoenfeld, Andreas A; Poppinga, Daniela; Harder, Dietrich; Doerner, Karl-Joachim; Poppe, Bjoern

    2014-07-07

    Optical experiments and theoretical considerations have been undertaken in order to understand the causes of the 'orientation effect' and the 'parabola effect', the artefacts impairing the desired light absorption measurement on radiochromic EBT3 films with flatbed scanners. EBT3 films exposed to doses up to 20.9 Gy were scanned with an Epson Expression 10000XL flatbed scanner in landscape and portrait orientation. The horizontally and vertically polarized light components of the scanner were determined, and another Epson Expression 10000XL flatbed scanner was disassembled to examine its optical components. The optical properties of exposed and unexposed EBT3 films were studied with incident polarized and unpolarized white light, and the transmitted red light was investigated for its polarization and scattering properties including the distribution of the scattering angles. Neutral density filters were studied for comparison. Guidance was sought from the theory of light scattering from rod-like macromolecular structures. The drastic dose-dependent variation of the transmitted total light current as function of the orientation of front and rear polarizers, interpreted by light scattering theory, shows that the radiation-induced polymerization of the monomers of EBT3 films produces light scattering oscillators preferably polarized at right angles with the coating direction of the film. The directional distribution of the scattered light is partly anisotropic, with a preferred scattering plane at right angles with the coating direction, indicating light scattering from stacks of coherently vibrating oscillators piled up along the monomer crystals. The polyester carrier film also participates in these effects. The 'orientation' and 'parabola' artefacts due to flatbed scanning of radiochromic films can be explained by the interaction of the polarization-dependent and anisotropic light scattering from exposed and unexposed EBT3 films with the quantitative difference

  9. Radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Soares, Christopher G.

    2006-01-01

    The object of this paper is to give a new user some practical information on the use of radiochromic films for medical applications. While various aspects of radiochromic film dosimetry for medical applications have been covered in some detail in several other excellent review articles which have appeared in the last few years [Niroomand-Rad, A., Blackwell, C.R., Coursey, B.M., Gall, K.P., McLaughlin, W.L., Meigooni, A.S., Nath, R., Rodgers, J.E., Soares, C.G., 1998. Radiochromic dosimetry: recommendations of the AAPM Radiation Therapy Committee Task Group 55. Med. Phys. 25, 2093-2115; Dempsey, J.F., Low, D.A., Mutic, S., Markman, J., Kirov, A.S., Nussbaum, G.H., Williamson, J.F., 2000. Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions. Med. Phys. 27, 2462-2475; Butson, M.J., Yu, P.K.N., Cheung, T., Metcalfe, P., 2003. Radiochromic film for medical radiation dosimetry. Mater. Sci. Eng. R41, 61-120], it is the intent of the present author to present material from a more user-oriented and practical standpoint. That is, how the films work will be stressed much less than how to make the films work well. The strength of radiochromic films is most evident in applications where there is a very high dose gradient and relatively high absorbed dose rates. These conditions are associated with brachytherapy applications, measurement of small fields, and at the edges (penumbra regions) of larger fields

  10. Measurement of high energy x-ray beam penumbra with Gafchromic trade mark sign EBT radiochromic film

    International Nuclear Information System (INIS)

    Cheung Tsang; Butson, Martin J.; Yu, Peter K. N.

    2006-01-01

    High energy x-ray beam penumbra are measured using Gafchromic trade mark sign EBT film. Gafchromic trade mark sign EBT, due to its limited energy dependence and high spatial resolution provide a high level of accuracy for dose assessment in penumbral regions. The spatial resolution of film detector systems is normally limited by the scanning resolution of the densitometer. Penumbral widths (80%/20%) measured at D max were found to be 2.8, 3.0, 3.2, and 3.4 mm (±0.2 mm) using 5, 10, 20, and 30 cm square field sizes, respectively, for a 6 MV linear accelerator produced x-ray beam. This is compared to 3.2 mm±0.2 mm (Kodak EDR2) and 3.6 mm±0.2 mm (Kodak X-Omat V) at 10 cmx10 cm measured using radiographic film. Using a zero volume extrapolation technique for ionization chamber measurements, the 10 cmx10 cm field penumbra at D max was measured to be 3.1 mm, a close match to Gafchromic trade mark sign EBT results. Penumbral measurements can also be made at other depths, including the surface, as the film does not suffer significantly from dosimetric variations caused by changing x-ray energy spectra. Gafchromic trade mark sign EBT film provides an adequate measure of penumbral dose for high energy x-ray beams

  11. Radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Xu Zhiyong

    2002-01-01

    Radiochromic film dosimetry was developed to measure ionization irradiation dose for industry and medicine. At this time, there are no comprehensive guideline on the medical application, calibration method and densitometer system for medicine. The review gives update on Radiochromic film dosimetry used for medicine, including principles, film model and material, characteristics, calibration method, scanning densitometer system and medical application

  12. Evaluation of material heterogeneity dosimetric effects using radiochromic film for COMS eye plaques loaded with 125I seeds (model I25.S16)

    International Nuclear Information System (INIS)

    Acar, Hilal; Chiu-Tsao, Sou-Tung; Özbay, İsmail; Kemikler, Gönül; Tuncer, Samuray

    2013-01-01

    Purpose: (1) To measure absolute dose distributions in eye phantom for COMS eye plaques with 125 I seeds (model I25.S16) using radiochromic EBT film dosimetry. (2) To determine the dose correction function for calculations involving the TG-43 formalism to account for the presence of the COMS eye plaque using Monte Carlo (MC) method specific to this seed model. (3) To test the heterogeneous dose calculation accuracy of the new version of Plaque Simulator (v5.3.9) against the EBT film data for this seed model. Methods: Using EBT film, absolute doses were measured for 125 I seeds (model I25.S16) in COMS eye plaques (1) along the plaque's central axis for (a) uniformly loaded plaques (14–20 mm in diameter) and (b) a 20 mm plaque with single seed, and (2) in off-axis direction at depths of 5 and 12 mm for all four plaque sizes. The EBT film calibration was performed at 125 I photon energy. MC calculations using MCNP5 code for a single seed at the center of a 20 mm plaque in homogeneous water and polystyrene medium were performed. The heterogeneity dose correction function was determined from the MC calculations. These function values at various depths were entered into PS software (v5.3.9) to calculate the heterogeneous dose distributions for the uniformly loaded plaques (of all four sizes). The dose distributions with homogeneous water assumptions were also calculated using PS for comparison. The EBT film measured absolute dose rate values (film) were compared with those calculated using PS with homogeneous assumption (PS Homo) and heterogeneity correction (PS Hetero). The values of dose ratio (film/PS Homo) and (film/PS Hetero) were obtained. Results: The central axis depth dose rate values for a single seed in 20 mm plaque measured using EBT film and calculated with MCNP5 code (both in ploystyrene phantom) were compared, and agreement within 9% was found. The dose ratio (film/PS Homo) values were substantially lower than unity (mostly between 0.8 and 0.9) for all

  13. Simplified method for creating a density-absorbed dose calibration curve for the low dose range from Gafchromic EBT3 film

    Directory of Open Access Journals (Sweden)

    Tatsuhiro Gotanda

    2016-01-01

    Full Text Available Radiochromic film dosimeters have a disadvantage in comparison with an ionization chamber in that the dosimetry process is time-consuming for creating a density-absorbed dose calibration curve. The purpose of this study was the development of a simplified method of creating a density-absorbed dose calibration curve from radiochromic film within a short time. This simplified method was performed using Gafchromic EBT3 film with a low energy dependence and step-shaped Al filter. The simplified method was compared with the standard method. The density-absorbed dose calibration curves created using the simplified and standard methods exhibited approximately similar straight lines, and the gradients of the density-absorbed dose calibration curves were −32.336 and −33.746, respectively. The simplified method can obtain calibration curves within a much shorter time compared to the standard method. It is considered that the simplified method for EBT3 film offers a more time-efficient means of determining the density-absorbed dose calibration curve within a low absorbed dose range such as the diagnostic range.

  14. Characterization of beta radiation fields using radiochromic films; Caracterizacao de campos de radiacao beta utilizando filmes radiocromicos

    Energy Technology Data Exchange (ETDEWEB)

    Benavente, Jhonny A.; Silva, Teogenes A. da, E-mail: jabc@cdtn.b [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil). Programa de Pos-Graduacao em Ciencia e Tecnologia das Radiacoes, Minerais e Materiais; Meira-Belo, Luiz C.; Reynaldo, Sibele R. [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2011-07-01

    The objective of this work was to study the response of radiochromic films for beta radiation fields in terms of absorbed dose. The reliability of the EBT model Gafchromic radiochromic film was studied. A 9800 XL model Microtek, transmission scanner, a 369 model X-Rite optical densitometer and a Mini 1240 Shimadzu UV spectrophotometer were used for measurement comparisons. Calibration of the three systems was done with irradiated samples of radiochromic films with 0.1; 0.3; 0.5; 0.8; 1.0; 1.5; 2.0; 2.5; 3.0; 3.5; 4.5 e 5.0 Gy in beta radiation field from a Sr-90/Y-90 source. Calibration was performed by establishing a correlation between the absorbed dose values and the corresponding radiochromic responses. Results showed significant differences in the absorbed dose values obtained with the three methods. Absorbed dose values showed errors from 0.6 to 4.4%, 0.3 to 31.8% and 0.2 to 47.3% for the Microtek scanner, the X-Rite Densitometer and the Shimadzu spectrophotometer, respectively. Due to the easy acquisition and use for absorbed dose measurements, the densitometer and the spectrophotometer showed to be suitable techniques to evaluate radiation dose in relatively homogeneous fields. In the case of inhomogeneous fields or for a two dimension mapping of radiation fields to identify anisotropies, the scanner technique is the most recommended. (author)

  15. The artefacts of radiochromic film dosimetry with flatbed scanners and their causation by light scattering from radiation-induced polymers

    International Nuclear Information System (INIS)

    Schoenfeld, Andreas A; Poppinga, Daniela; Poppe, Bjoern; Harder, Dietrich; Doerner, Karl-Joachim

    2014-01-01

    Optical experiments and theoretical considerations have been undertaken in order to understand the causes of the ‘orientation effect’ and the ‘parabola effect’, the artefacts impairing the desired light absorption measurement on radiochromic EBT3 films with flatbed scanners. EBT3 films exposed to doses up to 20.9 Gy were scanned with an Epson Expression 10000XL flatbed scanner in landscape and portrait orientation. The horizontally and vertically polarized light components of the scanner were determined, and another Epson Expression 10000XL flatbed scanner was disassembled to examine its optical components. The optical properties of exposed and unexposed EBT3 films were studied with incident polarized and unpolarized white light, and the transmitted red light was investigated for its polarization and scattering properties including the distribution of the scattering angles. Neutral density filters were studied for comparison. Guidance was sought from the theory of light scattering from rod-like macromolecular structures. The drastic dose-dependent variation of the transmitted total light current as function of the orientation of front and rear polarizers, interpreted by light scattering theory, shows that the radiation-induced polymerization of the monomers of EBT3 films produces light scattering oscillators preferably polarized at right angles with the coating direction of the film. The directional distribution of the scattered light is partly anisotropic, with a preferred scattering plane at right angles with the coating direction, indicating light scattering from stacks of coherently vibrating oscillators piled up along the monomer crystals. The polyester carrier film also participates in these effects. The ‘orientation’ and ‘parabola’ artefacts due to flatbed scanning of radiochromic films can be explained by the interaction of the polarization-dependent and anisotropic light scattering from exposed and unexposed EBT3 films with the quantitative

  16. SU-E-T-37: An Optical Investigation Into the Polarization and Scattering Effects Underlying the Artifacts of Radiochromic Film Dosimetry with Commercial Flatbed Scanners

    International Nuclear Information System (INIS)

    Schoenfeld, A; Poppinga, D; Poppe, B; Harder, D; Doerner, K

    2014-01-01

    Purpose: This study aims to investigate the optical properties of radiochromic EBT3 films on exposure to polarized incident light. Methods: An optical table setup was used to investigate the properties of exposed and unexposed EBT3 films. The films were placed with their long side horizontally and illuminated with polarized incident white light. The polarization of light with the electrical vector pointing vertically is referred to as 0°, accordingly horizontal orientation corresponds to 90°. The light transmission was measured depending on the polarization angle of the incident light and the polarization of a polarizer in front of the detector. Secondly, the scattering properties of exposed and unexposed films were measured by placing a plane convex lens behind the films and a screen in its focal plane. Thereby, the distribution of the scattering angles appears as an intensity map on the screen. The distributions of scattering angles caused by EBT3 films and by neutral density filters were compared. Results: EBT3 films show a strong dependence of the light transmission on the polarization of the incident light. With both polarizers parallel, a peak transmission was found at 90° orientation of the polarizers. With the rear polarizer at right angles with the front polarizer, peak transmissions were found at front polarizer orientations 45° and 135°. The scattering appears to be anisotropic with a preference direction parallel to the long side of the film. The portion of scattered light and the half value scattering angle both increase with the dose on the film. Conclusion: EBT3 films show dose dependent changes in polarized light transmission and anisotropic light scattering. These effects impair the light absorption measurements on exposed films performed with commercial flatbed scanners and are causing the well-known artifacts of radiochromic film dosimetry with flatbed scanners, the “orientation effect” and the “parabola effect”

  17. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic

    International Nuclear Information System (INIS)

    Palmer, A. L.; Di Pietro, P.; Alobaidli, S.; Issa, F.; Doran, S.; Bradley, D.; Nisbet, A.

    2013-01-01

    Purpose: Dose distribution measurement in clinical high dose rate (HDR) brachytherapy is challenging, because of the high dose gradients, large dose variations, and small scale, but it is essential to verify accurate treatment planning and treatment equipment performance. The authors compare and evaluate three dosimetry systems for potential use in brachytherapy dose distribution measurement: Ge-doped optical fibers, EBT3 Gafchromic film with multichannel analysis, and the radiochromic material PRESAGE ® with optical-CT readout. Methods: Ge-doped SiO 2 fibers with 6 μm active core and 5.0 mm length were sensitivity-batched and their thermoluminescent properties used via conventional heating and annealing cycles. EBT3 Gafchromic film of 30 μm active thickness was calibrated in three color channels using a nominal 6 MV linear accelerator. A 48-bit transmission scanner and advanced multichannel analysis method were utilized to derive dose measurements. Samples of the solid radiochromic polymer PRESAGE ® , 60 mm diameter and 100 mm height, were analyzed with a parallel beam optical CT scanner. Each dosimetry system was used to measure the dose as a function of radial distance from a Co-60 HDR source, with results compared to Monte Carlo TG-43 model data. Each system was then used to measure the dose distribution along one or more lines through typical clinical dose distributions for cervix brachytherapy, with results compared to treatment planning system (TPS) calculations. Purpose-designed test objects constructed of Solid Water and held within a full-scatter water tank were utilized. Results: All three dosimetry systems reproduced the general shape of the isolated source radial dose function and the TPS dose distribution. However, the dynamic range of EBT3 exceeded those of doped optical fibers and PRESAGE ® , and the latter two suffered from unacceptable noise and artifact. For the experimental conditions used in this study, the useful range from an isolated

  18. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic.

    Science.gov (United States)

    Palmer, A L; Di Pietro, P; Alobaidli, S; Issa, F; Doran, S; Bradley, D; Nisbet, A

    2013-06-01

    Dose distribution measurement in clinical high dose rate (HDR) brachytherapy is challenging, because of the high dose gradients, large dose variations, and small scale, but it is essential to verify accurate treatment planning and treatment equipment performance. The authors compare and evaluate three dosimetry systems for potential use in brachytherapy dose distribution measurement: Ge-doped optical fibers, EBT3 Gafchromic film with multichannel analysis, and the radiochromic material PRESAGE(®) with optical-CT readout. Ge-doped SiO2 fibers with 6 μm active core and 5.0 mm length were sensitivity-batched and their thermoluminescent properties used via conventional heating and annealing cycles. EBT3 Gafchromic film of 30 μm active thickness was calibrated in three color channels using a nominal 6 MV linear accelerator. A 48-bit transmission scanner and advanced multichannel analysis method were utilized to derive dose measurements. Samples of the solid radiochromic polymer PRESAGE(®), 60 mm diameter and 100 mm height, were analyzed with a parallel beam optical CT scanner. Each dosimetry system was used to measure the dose as a function of radial distance from a Co-60 HDR source, with results compared to Monte Carlo TG-43 model data. Each system was then used to measure the dose distribution along one or more lines through typical clinical dose distributions for cervix brachytherapy, with results compared to treatment planning system (TPS) calculations. Purpose-designed test objects constructed of Solid Water and held within a full-scatter water tank were utilized. All three dosimetry systems reproduced the general shape of the isolated source radial dose function and the TPS dose distribution. However, the dynamic range of EBT3 exceeded those of doped optical fibers and PRESAGE(®), and the latter two suffered from unacceptable noise and artifact. For the experimental conditions used in this study, the useful range from an isolated HDR source was 5-40 mm for

  19. Radiochromic film and polarization effects

    International Nuclear Information System (INIS)

    Yu, P.K.N.; Cheung, T.; Butson, M.J.; Cancer Services, Wollongong, NSW; Inwood, D.

    2004-01-01

    Full text: A new high sensitivity radiochromic film has been tested for its polarization properties. Gafchromic HS film has been shown to produce a relatively small (less than 3%) variation in measured optical density measured at 660nm wavelength when the light source is fully linear polarized and the film is rotated through 360 deg angle. Similar variations are seen when the detector is linearly polarized. If both light source and detector is linearly polarised variations in measured optical density can reach 15% when the film is rotated through 360 deg angle. This seems to be due to a phase shift in polarised light caused by the radiochromic film resulting in the polarised light source becoming out of phase with the polarised detector. Gafchromic HS radiochromic film produces a minimal polarization response with varying angle of rotation however we recommend that a polarization test be performed on a densitometry system to establish the extent of its polarization properties before accuracy dosimetry is performed with radiochromic HS film. Copyright (2004) Australasian College of Physical Scientists and Engineers in Medicine

  20. A spectroscopic study of the chromatic properties of GafChromic™EBT3 films

    Energy Technology Data Exchange (ETDEWEB)

    Callens, M., E-mail: maarten.callens@kuleuven-kulak.be; Van Den Abeele, K. [Wave Propagation and Signal Processing, KU Leuven–KULAK, Kortrijk 8500 (Belgium); Crijns, W.; Depuydt, T.; Haustermans, K. [Department of Radiation Oncology, University Hospitals Leuven, Leuven 3000 (Belgium); Simons, V. [imec, Kapeldreef 75, Leuven 3001 (Belgium); De Wolf, I. [imec, Kapeldreef 75, Leuven 3001, Belgium and Department of Materials Engineering, KU Leuven, Leuven 3001 (Belgium); Maes, F. [Department of Electrical Engineering, ESAT/PSI, KU Leuven, Leuven 3001 (Belgium); D’hooge, J. [Department of Cardiovascular Sciences, KU Leuven, Leuven 3000 (Belgium); D’Agostino, E. [DoseVue NV, Hasselt 3500 (Belgium); Wevers, M.; Pfeiffer, H. [Department of Materials Engineering, KU Leuven, Leuven 3001 (Belgium)

    2016-03-15

    Purpose: This work provides an interpretation of the chromatic properties of GafChromic™EBT3 films based on the chemical nature of the polydiacetylene (PDA) molecules formed upon interaction with ionizing radiation. The EBT3 films become optically less transparent with increasing radiation dose as a result of the radiation-induced polymerization of diacetylene monomers. In contrast to empirical quantification of the chromatic properties, less attention has been given to the underlying molecular mechanism that induces the strong decrease in transparency. Methods: Unlaminated GafChromic™EBT3 films were irradiated with a 6 MV photon beam to dose levels up to 20 Gy. The optical absorption properties of the films were investigated using visible (vis) spectroscopy. The presence of PDA molecules in the active layer of the EBT3 films was investigated using Raman spectroscopy, which probes the vibrational modes of the molecules in the layer. The vibrational modes assigned to PDA’s were used in a theoretical vis-absorption model to fit our experimental vis-absorption spectra. From the fit parameters, one can assess the relative contribution of different PDA conformations and the length distribution of PDA’s in the film. Results: Vis-spectroscopy shows that the optical density increases with dose in the full region of the visible spectrum. The Raman spectrum is dominated by two vibrational modes, most notably by the ν(C≡C) and the ν(C=C) stretching modes of the PDA backbone. By fitting the vis-absorption model to experimental spectra, it is found that the active layer contains two distinct PDA conformations with different absorption properties and reaction kinetics. Furthermore, the mean PDA conjugation length is found to be 2–3 orders of magnitude smaller than the crystals PDA’s are embedded in. Conclusions: Vis- and Raman spectroscopy provided more insight into the molecular nature of the radiochromic properties of EBT3 films through the identification of

  1. Using Radiochromic Films to Characterize the Dispersion of ZrO{sub 2} Nano-sized Grain Clusters in Protective Polymer Composites

    Energy Technology Data Exchange (ETDEWEB)

    Fontainha, C.C.P.; Nolasco, A.V. [Depto. de Engenharia Nuclear - DEN / UFMG - MG, Av. Antonio Carlos 6627, 31270-970 Belo Horizonte, MG (Brazil); Santos, A.P.; Faria, L.O. [Centro de Desenvolvimento da Tecnologia Nuclear, Av. Antonio Carlos 6627, C.P. 941, 30270-901, Belo Horizonte, MG (Brazil)

    2015-07-01

    Radiation dosimetry is commonly used to prevent deterministic radiation effects in high dose medical procedures. Radiochromic films find nowadays widely application in radiotherapy, interventional procedures and CT exams for isodose and maximum skin dose measurements. Moreover the size of the irradiated area and its distribution can be performed through the reading of the individual components in the RGB-spectrum. Particularly, radiochromic film has multiple advantages over alternative dosimeters for low-kV X-rays dosimetry. Concerned to spatial resolution it is far superior to that of ionization chambers and thermoluminescent dosimeters. For high energy photon fields (keV to MeV) the most used radiochromic film commercially available belongs to the EBT Gafchromic{sup R} series. On the other hand, for low energy photon fields in the x-ray range (20 kVp to 200 kVp) the best choice belongs to the XR-QA Gafchromic{sup R} film series. In this work we demonstrate the possibility of generating 2D images of thin polymeric composites films using EBT3 and XR-QA2 Gafchromic{sup R} films exposed to 6 MeV and 40 keV x-ray photons, respectively, using the digital filtering tools of the ImageJ{sup R} free software. In this context, EBT3 films were placed on the surface of a rigid anthropomorphic phantom. Then, they were covered with a thermoplastic mask made of PCL polymer. This setup was then exposed to 2.0 Gy absorbed dose in the Linear Accelerator beam. The EBT3 films were then scanned in the high resolution mode in a commercial scanner and the images subsequently treated with digital filters. It is somehow possible to see the image of the thermoplastic mask in the scanned image. However, in the treated image it is easy to observe the mask arrangement. The unexpected phenomenon here is the EB3 film ability to detect the attenuation of high energy photons by a plastic material, which in turn has a very low mass-energy attenuation coefficient, producing a very clear 2D image

  2. Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system.

    Science.gov (United States)

    Aldelaijan, Saad; Devic, Slobodan

    2018-05-01

    Different dose response functions of EBT3 model GafChromic™ film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation. Pieces of EBT3 film model were irradiated to different dose values in Solid Water (SW) phantom. Based on images scanned in both reflection and transmission mode before and after irradiation, twelve different response functions were calculated. For every response function, a reference radiochromic film dosimetry system was established by generating calibration curve and by performing the error vs. uncertainty analysis. Response functions using pixel values from the green channel demonstrated the highest sensitivity in both transmission and reflection mode. All functions were successfully fitted with rational functional form, and provided an overall one-sigma uncertainty of better than 2% for doses above 2 Gy. Use of pre-scanned images to calculate response functions resulted in negligible improvement in dose measurement accuracy. Although reflection scanning mode provides higher sensitivity and could lead to a more widespread use of radiochromic film dosimetry, it has fairly limited dose range and slightly increased uncertainty when compared to transmission scan based response functions. Double-scanning technique, either in transmission or reflection mode, shows negligible improvement in dose accuracy as well as a negligible increase in dose uncertainty. Normalized pixel value of the images scanned in transmission mode shows linear response in a dose range of up to 11 Gy. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  3. Design of a protocol for the use of radiochromic films in IMRT plans control

    International Nuclear Information System (INIS)

    Aberbuj, P.D.

    2011-01-01

    The purpose of this work is to design a protocol for the use of the Gafchromic EBT2 radiochromic films with the Epson CX5600 scanner as dosimetric system for IMRT patient specific quality assurance, with an emphasis on keep the uncertainty below 3%. The studied uncertainty sources are related to the scanner reproducibility, the film and scanner homogeneity, and the dose estimation. By introducing a series of modifications to the initial protocol, the total uncertainty was kept below 3% in the range 30-500 c Gy, being less than 1% between 150 and 500 c Gy. (author)

  4. Radiochromic film calibration wedge EBT2 using virtual fields; Calibracion de peliculas radiocromicos EBT2 mediante campos con cunas virtuales

    Energy Technology Data Exchange (ETDEWEB)

    Perez, M. A.; Macias, J.; Merchan, M. A.; Campo, J. L.; Moreno, J. C.; Terron, J. A.; Miras, H.; Ortiz, M.; Arrans, R.; Ortiz, A.; Fernandez, D.

    2011-07-01

    EBT2 film dosimetry after exposure to a gradient of these wedge dosimetry. In our case a virtual wedge 600. The primary objective is to automate the process, reduce the time spent in obtaining the calibration curve (color-dose). Time negligible due to the limited availability of accelerators. This method of obtaining the calibration curve provides similar results to the commonly accepted either with irradiation uniform of a single film with different dose levels (multiband ladder) or with irradiation uniform of small rectangular piece of film , decreasing by a factor about 20, the time spent. (Author)

  5. SU-F-T-550: Radiochromic Plastic Thin Sheet Dosimeter: Initial Performance

    Energy Technology Data Exchange (ETDEWEB)

    Jordan, K [London Regional Cancer Program, London, ON (United Kingdom); Adamovics, J [John Adamovics, Skillman, NJ (United States)

    2016-06-15

    Purpose: Thin sheets, of a high sensitivity formulation of radiochromic dosimeter, Presage were prepared and evaluated for optical readout. Methods: Sheets of radiochromic polyurethane, 12 cm long, 10 cm wide and 0.2 cm thick were prepared with leuco crystal violet as the reporter molecule. Sample transmission was evaluated at a wavelength of 590 nm with in-house constructed instruments: optical cone beam laser CT scanner, fixed and scanning spot densitometers. Sample sequential irradiations to a total dose of 40 Gy were conducted with a modified, Theratron 60, cobalt radiotherapy machine at dose rates of 1 or 0.25 Gy per minute. Exposure to ambient and readout light was minimized to limit background photochromic signals. Samples were stored at 4°C. Optical activity was assessed from linearly polarized transmission images. Comparison sensitivity measurements with EBT3 film were conducted. Results: Samples were transparent, smooth and pale purple before irradiation. Radiochromic reaction was completed in less than 5 minutes. A linear dose response with a sensitivity of 0.5 cm-1Gy-1 was observed. Micrometer measurements found sheet thickness variations up to 20%. Uniform dose, 2 Gy attenuation images, correlated with local sheet thicknesses. Comparable measurements with EBT3 film were 3 times more sensitive at 1 Gy but above 15 Gy, EBT3 film had lower sensitivity than 0.2 cm thick Presage sheet dosimeter due to its non-linear response. Conclusion: Dose sensitivity provided a 10% decrease in transmission for a 1 Gy dose. Improvements in mold design are expected to allow production of sheets with less than 5% variation in thickness. Above, 10 Gy, Presage sheet dosimeter performance expected to exceed EBT3 film based on linearity, sensitivity, transparency and smoothness of samples. J Adamovics is owner of Heuris Inc.

  6. Medical radiation dosimetry with radiochromic film

    International Nuclear Information System (INIS)

    Butson, M.J.; Cancer Services, NSW; Cheung, T.; Yu, P.K.N.; Metcalfe, P.

    2004-01-01

    Full text: Photon, electron and proton radiation are used extensively for medical purposes in diagnostic and therapeutic procedures. Dosimetry of these radiation sources can be performed with radiochromic films, devices that have the ability to produce a permanent visible colour change upon irradiation. Within the last ten years, the use of radiochromic films has expanded rapidly in the medical world due to commercial products becoming more readily available, higher sensitivity films and technology advances in imaging which have allowed scientists to use two-dimensional dosimetry more accurately and inexpensively. Radiochromic film dosimeters are now available in formats, which have accurate dose measurement ranges from less than 1 Gy up to many kGy. A relatively energy independent dose response combined with automatic development of radiochromic film products has made these detectors most useful in medical radiation dosimetry. Copyright (2004) Australasian College of Physical Scientists and Engineers in Medicine

  7. Gafchromic EBT-XD film: Dosimetry characterization in high-dose, volumetric-modulated arc therapy.

    Science.gov (United States)

    Miura, Hideharu; Ozawa, Shuichi; Hosono, Fumika; Sumida, Naoki; Okazue, Toshiya; Yamada, Kiyoshi; Nagata, Yasushi

    2016-11-08

    Radiochromic films are important tools for assessing complex dose distributions. Gafchromic EBT-XD films have been designed for optimal performance in the 40-4,000 cGy dose range. We investigated the dosimetric characteristics of these films, including their dose-response, postexposure density growth, and dependence on scanner orientation, beam energy, and dose rate with applications to high-dose volumetric-modulated arc therapy (VMAT) verification. A 10 MV beam from a TrueBeam STx linear accelerator was used to irradiate the films with doses in the 0-4,000 cGy range. Postexposure coloration was analyzed at postirradiation times ranging from several minutes to 48 h. The films were also irradiated with 6 MV (dose rate (DR): 600 MU/min), 6 MV flattening filter-free (FFF) (DR: 1,400 MU/ min), and 10 MV FFF (DR: 2,400 MU/min) beams to determine the energy and dose-rate dependence. For clinical examinations, we compared the dose distribu-tion measured with EBT-XD films and calculated by the planning system for four VMAT cases. The red channel of the EBT-XD film exhibited a wider dynamic range than the green and blue channels. Scanner orientation yielded a variation of ~ 3% in the net optical density (OD). The difference between the film front and back scan orientations was negligible, with variation of ~ 1.3% in the net OD. The net OD increased sharply within the first 6 hrs after irradiation and gradually afterwards. No significant difference was observed for the beam energy and dose rate, with a variation of ~ 1.5% in the net OD. The gamma passing rates (at 3%, 3 mm) between the film- measured and treatment planning system (TPS)-calculated dose distributions under a high dose VMAT plan in the absolute dose mode were more than 98.9%. © 2016 The Authors.

  8. Dosimetry using radiochromic film and planning algorithms in heterogeneous phantoms

    International Nuclear Information System (INIS)

    Leite, Vinicius Freitas

    2012-01-01

    This work analyzes, through the study of the interaction of electromagnetic radiation with matter, two schemes of heterogeneous phantoms schematised to simulate real cases of planning with different electronic densities through the Pencil Beam, Collapsed Cone and Analytical Anisotropic Algorithm algorithms and compare with measurements Of relative absorbed dose in an IBA CC13 ionization chamber and Gafchromic® EBT2 radiochromic film. Epichlorohydrin rubber and its compatibility in comparison with human bone has also been evaluated. The assembly of the heterogeneous phantoms was feasible and the results regarding the density and attenuation of the rubber presented consistent values. However, the study of PDPs in constructed phantoms showed a considerable percentage discrepancy between measurements and planning

  9. High dose per fraction dosimetry of small fields with Gafchromic EBT2 film

    International Nuclear Information System (INIS)

    Hardcastle, Nicholas; Basavatia, Amar; Bayliss, Adam; Tome, Wolfgang A.

    2011-01-01

    Purpose: Small field dosimetry is prone to uncertainties due to the lack of electronic equilibrium and the use of the correct detector size relative to the field size measured. It also exhibits higher sensitivity to setup errors as well as large variation in output with field size and shape. Radiochromic film is an attractive method for reference dosimetry in small fields due to its ability to provide 2D dose measurements while having minimal impact on the dose distribution. Gafchromic EBT2 has a dose range of up to 40 Gy; therefore, it could potentially be useful for high dose reference dosimetry with high spatial resolution. This is a requirement in stereotactic radiosurgery deliveries, which deliver high doses per fraction to small targets. Methods: Targets of 4 mm and 12 mm diameters were treated to a minimum peripheral dose of 21 Gy prescribed to 80% of the maximum dose in one fraction. Target doses were measured with EBT2 film (both targets) and an ion chamber (12 mm target only). Measured doses were compared with planned dose distributions using profiles through the target and minimum peripheral dose coverage. Results: The measured target doses and isodose coverage agreed with the planned dose within ±1 standard deviation of three measurements, which were 2.13% and 2.5% for the 4 mm and 12 mm targets, respectively. Conclusions: EBT2 film is a feasible dosimeter for high dose per fraction reference 2D dosimetry.

  10. Fast protocol for radiochromic film dosimetry using a cloud computing web application.

    Science.gov (United States)

    Calvo-Ortega, Juan-Francisco; Pozo, Miquel; Moragues, Sandra; Casals, Joan

    2017-07-01

    To investigate the feasibility of a fast protocol for radiochromic film dosimetry to verify intensity-modulated radiotherapy (IMRT) plans. EBT3 film dosimetry was conducted in this study using the triple-channel method implemented in the cloud computing application (Radiochromic.com). We described a fast protocol for radiochromic film dosimetry to obtain measurement results within 1h. Ten IMRT plans were delivered to evaluate the feasibility of the fast protocol. The dose distribution of the verification film was derived at 15, 30, 45min using the fast protocol and also at 24h after completing the irradiation. The four dose maps obtained per plan were compared using global and local gamma index (5%/3mm) with the calculated one by the treatment planning system. Gamma passing rates obtained for 15, 30 and 45min post-exposure were compared with those obtained after 24h. Small differences respect to the 24h protocol were found in the gamma passing rates obtained for films digitized at 15min (global: 99.6%±0.9% vs. 99.7%±0.5%; local: 96.3%±3.4% vs. 96.3%±3.8%), at 30min (global: 99.5%±0.9% vs. 99.7%±0.5%; local: 96.5%±3.2% vs. 96.3±3.8%) and at 45min (global: 99.2%±1.5% vs. 99.7%±0.5%; local: 96.1%±3.8% vs. 96.3±3.8%). The fast protocol permits dosimetric results within 1h when IMRT plans are verified, with similar results as those reported by the standard 24h protocol. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  11. Note: Calibration of EBT3 radiochromic film for measuring solar ultraviolet radiation

    Energy Technology Data Exchange (ETDEWEB)

    Chun, S. L. [Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong (Hong Kong); Yu, P. K. N., E-mail: peter.yu@cityu.edu.hk [Department of Physics and Materials Science, City University of Hong Kong, Kowloon Tong (Hong Kong); State Key Laboratory in Marine Pollution, City University of Hong Kong, Kowloon Tong (Hong Kong)

    2014-10-01

    Solar (UVA + UVB) exposure was assessed using the Gafchromic EBT3 film. The coloration change was represented by the net reflective optical density (Net ROD). Through calibrations against a UV-tube lamp, operational relationships were obtained between Net ROD and the (UVA + UVB) exposures (in J cm⁻²p or J m⁻²). The useful range was from ~0.2 to ~30 J cm⁻². The uniformity of UV irradiation was crucial for an accurate calibration. For solar exposures ranging from 2 to 11 J cm⁻², the predicted Net ROD agreed with the recorded values within 9%, while the predicted exposures agreed with the recorded values within 15%.

  12. Dose rate dependence for different dosimeters and detectors: TLD, OSL, EBT films, and diamond detectors

    International Nuclear Information System (INIS)

    Karsch, L.; Beyreuther, E.; Burris-Mog, T.; Kraft, S.; Richter, C.; Zeil, K.; Pawelke, J.

    2012-01-01

    Purpose: The use of laser accelerators in radiation therapy can perhaps increase the low number of proton and ion therapy facilities in some years due to the low investment costs and small size. The laser-based acceleration technology leads to a very high peak dose rate of about 10 11 Gy/s. A first dosimetric task is the evaluation of dose rate dependence of clinical dosimeters and other detectors. Methods: The measurements were done at ELBE, a superconductive linear electron accelerator which generates electron pulses with 5 ps length at 20 MeV. The different dose rates are reached by adjusting the number of electrons in one beam pulse. Three clinical dosimeters (TLD, OSL, and EBT radiochromic films) were irradiated with four different dose rates and nearly the same dose. A faraday cup, an integrating current transformer, and an ionization chamber were used to control the particle flux on the dosimeters. Furthermore two diamond detectors were tested. Results: The dosimeters are dose rate independent up to 410 9 Gy/s within 2% (OSL and TLD) and up to 1510 9 Gy/s within 5% (EBT films). The diamond detectors show strong dose rate dependence. Conclusions: TLD, OSL dosimeters, and EBT films are suitable for pulsed beams with a very high pulse dose rate like laser accelerated particle beams.

  13. Evaluation of radiochromic films EBT3 for in-vivo dosimetry in radiotherapy treatments with photons; Evaluacion de peliculas radiocromicas EBT3 para dosimetria in vivo en tratamientos de radioterapia con fotones

    Energy Technology Data Exchange (ETDEWEB)

    Galvan de la C, O. O.; Rivera M, T. [IPN, Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Av. Legaria No. 694, Col. Irrigacion, 11500 Mexico D. F. (Mexico); Garcia G, O. A.; Larraga G, J. M., E-mail: olinca@ciencias.unam.mx [Instituto Nacional de Neurologia y Neurocirugia, Laboratorio de Fisica Medica, Insurgentes Sur 3877, Col. La Fama, 14269 Mexico D. F. (Mexico)

    2015-10-15

    Full text: In-vivo dosimetry is a challenge in radiotherapy due to the measures are carried out in reference conditions outside; there is no balance of charged particle and beam consists of photons own and contamination electrons. Detectors that are useful for such measures should be sufficiently small and thin so they do not disturb the beam and do not alter the dose on target. In this paper the radiochromic films Gafchromic model EBT3 are evaluated as potential detectors for in-vivo dosimetry; measurements were carried out in solid water phantom on the surface, with films of dimensions 3 x 3 cm{sup 2}. Irradiations were performed with a linear accelerator Novalis of 6 MV. Comparison between dose values found with a diode type Sfd detector (IBA dosimetry, Germany) and a diamond detector CVD (PTW-Freiburg) for 2 different sized of field (5 x 5 cm{sup 2} and 10 x 10 cm{sup 2}) on the surface of a water phantom scanning was realized. The total spreading factor (Tsf) measured on surface was of 0.831 ± 4.6%; which is greater 12.9% than Tsf measured at a depth of maximum dose. This difference may be due to the contribution of scattered electrons to the beam exit. The measures must be corroborated with Monte Carlo simulations, which they will be validated on surface by the Abdel-Rahman [et al.] method; this mechanism will determine if the films are useful detectors for in-vivo dosimetry clinically. (Author)

  14. In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment.

    Science.gov (United States)

    Severgnini, Mara; de Denaro, Mario; Bortul, Marina; Vidali, Cristiana; Beorchia, Aulo

    2014-01-08

    Intraoperative electron radiation therapy (IOERT) cannot usually benefit, as conventional external radiotherapy, from software systems of treatment planning based on computed tomography and from common dose verify procedures. For this reason, in vivo film dosimetry (IVFD) proves to be an effective methodology to evaluate the actual radiation dose delivered to the target. A practical method for IVFD during breast IOERT was carried out to improve information on the dose actually delivered to the tumor target and on the alignment of the shielding disk with respect to the electron beam. Two EBT3 GAFCHROMIC films have been positioned on the two sides of the shielding disk in order to obtain the dose maps at the target and beyond the disk. Moreover the postprocessing analysis of the dose distribution measured on the films provides a quantitative estimate of the misalignment between the collimator and the disk. EBT3 radiochromic films have been demonstrated to be suitable dosimeters for IVD due to their linear dose-optical density response in a narrow range around the prescribed dose, as well as their capability to be fixed to the shielding disk without giving any distortion in the dose distribution. Off-line analysis of the radiochromic film allowed absolute dose measurements and this is indeed a very important verification of the correct exposure to the target organ, as well as an estimate of the dose to the healthy tissue underlying the shielding. These dose maps allow surgeons and radiation oncologists to take advantage of qualitative and quantitative feedback for setting more accurate treatment strategies and further optimized procedures. The proper alignment using elastic bands has improved the absolute dose accuracy and the collimator disk alignment by more than 50%.

  15. A new implementation of multichannel radiochromic film dosimetry; Una nueva implementacion del analisis multicanal para la dosimetria mediante peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez Rodriguez, C.; Martin Martin, G.; Bermudez Luna, R.; Lopez Fernandez, A.; Tores Olombrada, M. V. de

    2014-07-01

    The aims of this paper are to carry out a new implementation of the multichannel radiochromic film dosimetry (Micke A, Lewis D, Yu X. Multichannel film dosimetry with nonuniformity correction. Med Phys 2011;38:2523-34), to quantify the variation in gamma index as compared to the single channel film dosimetry, and to determine if the procedure achieves similar results by means of a different scanner that the one used by Micke et al. Radiochromic EBT2 films and a Microtek 9000 XL scanner were used. Our procedure simplifies the system calibration splitting it into two factors, manufactured batch and digitalization specific. Absorbed dose spatial distributions from an open radiotherapy beam without any modulation and 20 IMRT treatments were determined. Their gamma index maps were calculated and a comparison of the results from single channel and multichannel dosimetry was performed. A 5% mean increase in concordance was obtained by using the multichannel film dosimetry. Our results are similar to those reported by Micke et al. even though we are using a different scanner. (Author)

  16. A practical procedure to improve the accuracy of radiochromic film dosimetry. A integration with a correction method of uniformity correction and a red/blue correction method

    International Nuclear Information System (INIS)

    Uehara, Ryuzo; Tachibana, Hidenobu; Ito, Yasushi; Yoshino, Shinichi; Matsubayashi, Fumiyasu; Sato, Tomoharu

    2013-01-01

    It has been reported that the light scattering could worsen the accuracy of dose distribution measurement using a radiochromic film. The purpose of this study was to investigate the accuracy of two different films, EDR2 and EBT2, as film dosimetry tools. The effectiveness of a correction method for the non-uniformity caused from EBT2 film and the light scattering was also evaluated. In addition the efficacy of this correction method integrated with the red/blue correction method was assessed. EDR2 and EBT2 films were read using a flatbed charge-coupled device scanner (EPSON 10000 G). Dose differences on the axis perpendicular to the scanner lamp movement axis were within 1% with EDR2, but exceeded 3% (Maximum: +8%) with EBT2. The non-uniformity correction method, after a single film exposure, was applied to the readout of the films. A corrected dose distribution data was subsequently created. The correction method showed more than 10%-better pass ratios in dose difference evaluation than when the correction method was not applied. The red/blue correction method resulted in 5%-improvement compared with the standard procedure that employed red color only. The correction method with EBT2 proved to be able to rapidly correct non-uniformity, and has potential for routine clinical intensity modulated radiation therapy (IMRT) dose verification if the accuracy of EBT2 is required to be similar to that of EDR2. The use of red/blue correction method may improve the accuracy, but we recommend we should use the red/blue correction method carefully and understand the characteristics of EBT2 for red color only and the red/blue correction method. (author)

  17. [A practical procedure to improve the accuracy of radiochromic film dosimetry: a integration with a correction method of uniformity correction and a red/blue correction method].

    Science.gov (United States)

    Uehara, Ryuzo; Tachibana, Hidenobu; Ito, Yasushi; Yoshino, Shinichi; Matsubayashi, Fumiyasu; Sato, Tomoharu

    2013-06-01

    It has been reported that the light scattering could worsen the accuracy of dose distribution measurement using a radiochromic film. The purpose of this study was to investigate the accuracy of two different films, EDR2 and EBT2, as film dosimetry tools. The effectiveness of a correction method for the non-uniformity caused from EBT2 film and the light scattering was also evaluated. In addition the efficacy of this correction method integrated with the red/blue correction method was assessed. EDR2 and EBT2 films were read using a flatbed charge-coupled device scanner (EPSON 10000G). Dose differences on the axis perpendicular to the scanner lamp movement axis were within 1% with EDR2, but exceeded 3% (Maximum: +8%) with EBT2. The non-uniformity correction method, after a single film exposure, was applied to the readout of the films. A corrected dose distribution data was subsequently created. The correction method showed more than 10%-better pass ratios in dose difference evaluation than when the correction method was not applied. The red/blue correction method resulted in 5%-improvement compared with the standard procedure that employed red color only. The correction method with EBT2 proved to be able to rapidly correct non-uniformity, and has potential for routine clinical IMRT dose verification if the accuracy of EBT2 is required to be similar to that of EDR2. The use of red/blue correction method may improve the accuracy, but we recommend we should use the red/blue correction method carefully and understand the characteristics of EBT2 for red color only and the red/blue correction method.

  18. Characterization of an extrapolation chamber and radiochromic films for verifying the metrological coherence among beta radiation fields

    International Nuclear Information System (INIS)

    Castillo, Jhonny Antonio Benavente

    2011-01-01

    The metrological coherence among standard systems is a requirement for assuring the reliability of dosimetric quantities measurements in ionizing radiation field. Scientific and technologic improvements happened in beta radiation metrology with the installment of the new beta secondary standard BSS2 in Brazil and with the adoption of the internationally recommended beta reference radiations. The Dosimeter Calibration Laboratory of the Development Center for Nuclear Technology (LCD/CDTN), in Belo Horizonte, implemented the BSS2 and methodologies are investigated for characterizing the beta radiation fields by determining the field homogeneity, the accuracy and uncertainties in the absorbed dose in air measurements. In this work, a methodology to be used for verifying the metrological coherence among beta radiation fields in standard systems was investigated; an extrapolation chamber and radiochromic films were used and measurements were done in terms of absorbed dose in air. The reliability of both the extrapolation chamber and the radiochromic film was confirmed and their calibrations were done in the LCD/CDTN in 90 Sr/ 90 Y, 85 Kr and 147 Pm beta radiation fields. The angular coefficients of the extrapolation curves were determined with the chamber; the field mapping and homogeneity were obtained from dose profiles and isodose with the radiochromic films. A preliminary comparison between the LCD/CDTN and the Instrument Calibration Laboratory of the Nuclear and Energy Research Institute / Sao Paulo (LCI/IPEN) was carried out. Results with the extrapolation chamber measurements showed in terms of absorbed dose in air rates showed differences between both laboratories up to de -I % e 3%, for 90 Sr/ 90 Y, 85 Kr and 147 Pm beta radiation fields, respectively. Results with the EBT radiochromic films for 0.1, 0.3 and 0.15 Gy absorbed dose in air, for the same beta radiation fields, showed differences up to 3%, -9% and -53%. The beta radiation field mappings with

  19. The origin of the flatbed scanner artifacts in radiochromic film dosimetry—key experiments and theoretical descriptions

    Science.gov (United States)

    Schoenfeld, Andreas A.; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-11-01

    The optical origin of the lateral response and orientation artifacts, which occur when using EBT3 and EBT-XD radiochromic films together with flatbed scanners, has been reinvestigated by experimental and theoretical means. The common feature of these artifacts is the well-known parabolic increase in the optical density OD(x)  =  -log10 I(x)/I 0(x) versus offset x from the scanner midline (Poppinga et al 2014 Med. Phys. 41 021707). This holds for landscape and portrait orientations as well as for the three color channels. Dose-independent optical subjects, such as neutral density filters, linear polarizers, the EBT polyester foil and diffusive glass, also present the parabolic lateral artifact when scanned with a flatbed scanner. The curvature parameter c of the parabola function OD(x)  =  c 0  +  cx 2 is found to be a linear function of the dose, the parameters of which are influenced by the film orientation and film type, EBT3 or EBT-XD. The ubiquitous parabolic shape of function OD(x) is attributed (a) to the optical path-length effect (van Battum et al 2016 Phys. Med. Biol. 61 625-49), due to the increasing obliquity of the optical scanner light associated with increasing offset x from the scanner midline, and (b) and (c) to the partial polarization and scattering of the light leaving the film, which affect the ratio ~I(x)/{{I}0}(x) , thus making OD(x) increase with x 2. The orientation effect results from the changes of effects (b) and (c) associated with turning the film position, and thereby the orientation of the polymer structure of the sensitive film layer. In a comparison of experimental results obtained with selected optical subjects, the relative weights of the contributions of the optical path-length effect and the polarization and scattering of light leaving the films to the lateral response artifact have been estimated to be of the same order of magnitude. Mathematical models of these causes for the parabolic shape of function

  20. The origin of the flatbed scanner artifacts in radiochromic film dosimetry-key experiments and theoretical descriptions.

    Science.gov (United States)

    Schoenfeld, Andreas A; Wieker, Soeren; Harder, Dietrich; Poppe, Bjoern

    2016-11-07

    The optical origin of the lateral response and orientation artifacts, which occur when using EBT3 and EBT-XD radiochromic films together with flatbed scanners, has been reinvestigated by experimental and theoretical means. The common feature of these artifacts is the well-known parabolic increase in the optical density OD(x)  =  -log 10 I(x)/I 0 (x) versus offset x from the scanner midline (Poppinga et al 2014 Med. Phys. 41 021707). This holds for landscape and portrait orientations as well as for the three color channels. Dose-independent optical subjects, such as neutral density filters, linear polarizers, the EBT polyester foil and diffusive glass, also present the parabolic lateral artifact when scanned with a flatbed scanner. The curvature parameter c of the parabola function OD(x)  =  c 0   +  cx 2 is found to be a linear function of the dose, the parameters of which are influenced by the film orientation and film type, EBT3 or EBT-XD. The ubiquitous parabolic shape of function OD(x) is attributed (a) to the optical path-length effect (van Battum et al 2016 Phys. Med. Biol. 61 625-49), due to the increasing obliquity of the optical scanner light associated with increasing offset x from the scanner midline, and (b) and (c) to the partial polarization and scattering of the light leaving the film, which affect the ratio [Formula: see text], thus making OD(x) increase with x 2 . The orientation effect results from the changes of effects (b) and (c) associated with turning the film position, and thereby the orientation of the polymer structure of the sensitive film layer. In a comparison of experimental results obtained with selected optical subjects, the relative weights of the contributions of the optical path-length effect and the polarization and scattering of light leaving the films to the lateral response artifact have been estimated to be of the same order of magnitude. Mathematical models of these causes for the parabolic shape of

  1. Commissioning a hobby cutting device for radiochromic film preparation.

    Science.gov (United States)

    Zolfaghari, Somayeh; Francis, Kirby E; Kairn, Tanya; Crowe, Scott B

    2017-06-01

    In addition to a high spatial resolution and well characterised dose response, one of the major advantages of radiochromic film as a dosimeter is that sheets of film can be cut into pieces suitable for use as calibration films, and for in vivo and phantom measurements. The cutting of film is typically done using scissors or a guillotine, and this process can be time-consuming, limited in precision, requires extensive handling and does not allow holes to be cut from the film without cutting from an existing edge. This study investigated the use of a Brother ScanNCut hobby cutting system for EBT3 film preparation. The optimal operating parameters (blade size, pressure, speed) that resulted in precise cuts with minimal delamination at cut edges were identified using test cutting patterns. These parameters were then used to cut a large film insert for a stereotactic head phantom for comparison against an insert cut with scissors. While the hobby cutting system caused a wider region of delamination at the film edge (1.8 mm) compared to scissors (1 mm), the hobby cutting system was found to be able to produce reproducible cuts more efficiently and more accurately than scissors. The use of the hobby cutting system is recommended for complex phantom inserts (containing sharp corners or holes for alignment rods) or in situations where large numbers of film pieces need to be prepared.

  2. Improving radiochromic film's sensitivity by wrapping it with UV phosphor

    International Nuclear Information System (INIS)

    Geso, Moshi; Ackerly, Trevor; Patterson, William

    2004-01-01

    The main advantage of radiochromic-film dosimeters is the coupling of rapid full planar-acquisition, high-spatial resolution and dose linearity. Their main limitation, however, is their low radiation sensitivity. This precludes their application to measuring doses below a few Gy. Radiochromic films are sensitive to ultraviolet radiation. In this note the results of exposing radiochromic films to x rays through an UV imaging cassette, which converts and amplifies x rays to UV radiation, are presented. These results indicate a clear increase (around 50%) in radiochromic film's sensitivity to MV x rays (6, 10, and 18 MV) when exposed through the UV phosphor

  3. Technical Note: Initial characterization of the new EBT-XD Gafchromic film.

    Science.gov (United States)

    Grams, Michael P; Gustafson, Jon M; Long, Kenneth M; de los Santos, Luis E Fong

    2015-10-01

    To assess the dosimetric accuracy and energy dependence of the new EBT-eXtended Dose (XD) Gafchromic film and to compare the lateral response artifact (LRA) between EBT-XD and EBT3 film. EBT3 and EBT-XD calibration curves were created by exposing films to known doses from 0 to 3000 cGy using a 6 MV beam. To assess the accuracy and dynamic range of EBT-XD, a 60° enhanced dynamic wedge (EDW) was used to deliver a dose range of approximately 200-2900 cGy. Comparison to treatment planning system (TPS) calculation was made using a gamma analysis with 2%/2 mm passing criteria. To assess and compare the LRA between EBT3 and EBT-XD, 21 × 21 cm(2) open fields delivered doses of 1000, 2000, and 3000 cGy to both types of film. Films were placed at the center of the scanner, and ratios of measured to TPS predicted doses were calculated at 50 and 80 mm lateral from the scanner center in order to quantitatively assess the LRA. To evaluate the energy dependence of EBT-XD film, seven known doses ranging from 400 to 3000 cGy were delivered using both 6 and 18 MV beams and the resulting optical densities (ODs) compared. The gamma passing rate was 99.1% for the 6 MV EDW delivery. EBT-XD film exhibited minimal LRA (XD films exposed to known doses from 6 to 18 MV beams were XD is greatly reduced when compared to EBT3. This in combination with its accuracy from 0 to 3000 cGy and minimal energy dependence from 6 to 18 MV makes EBT-XD film well suited for dosimetric measurements in high dose SRS/SBRT applications.

  4. Verification and optimization of HDR surface mould brachytherapy plans using GAFCHROMIC EBT2 film: the ideal geometric case

    International Nuclear Information System (INIS)

    Sobolewski, Matthew; Haque, Mamoon

    2011-01-01

    Full text: Surface mould brachytherapy is used to treat superficial cancers due to conformal dose distributions and rapid dose fall-off with depth. In this work, we determine the effect of varying catheter number and prescription distance on dose distributions for surface mould plans using radiochromic film. Eight surface mould plans were generated using PLATO BPS (Version 14.3.2). Measurements were taken with Gafchromic EBT2 film over depths of 5-30 mm with an Ir-192 HDR source. Films were scanned using an Epson Expression 10000 XL flatbed scanner and analysed using RIT 113 software. The EBT2 films showed good agreement with an average difference of 2.8% compared to the planning system. The dose gradient in the interval ranging ±5 mm from the prescription point showed an 80% increase from the plan with maximum catheters (II) to the minimum (3). The size and extent of local dose maxima increased when fewer catheters were used. Increasing prescription distance decreased the dose gradient with a 20% reduction in dose occurring 4 mm superficially to the prescription point when prescription distance increased from 5 to 20 mm. Gafchromic EBT2 was used successfully to evaluate surface mould brachytherapy plans and is a useful tool for dose verification checks. High dose regions ne,u' to the catheter plane can be reduced by using a larger number of catheters and the prescription distance should be adjusted as a function of treatment depth varied by mould thickness.

  5. Technical Note: Initial characterization of the new EBT-XD Gafchromic film

    Energy Technology Data Exchange (ETDEWEB)

    Grams, Michael P., E-mail: grams.michael@mayo.edu; Gustafson, Jon M.; Long, Kenneth M.; Fong de los Santos, Luis E. [Department of Radiation Oncology, Mayo Clinic, Rochester, Minnesota 55905 (United States)

    2015-10-15

    Purpose: To assess the dosimetric accuracy and energy dependence of the new EBT-eXtended Dose (XD) Gafchromic film and to compare the lateral response artifact (LRA) between EBT-XD and EBT3 film. Methods: EBT3 and EBT-XD calibration curves were created by exposing films to known doses from 0 to 3000 cGy using a 6 MV beam. To assess the accuracy and dynamic range of EBT-XD, a 60° enhanced dynamic wedge (EDW) was used to deliver a dose range of approximately 200–2900 cGy. Comparison to treatment planning system (TPS) calculation was made using a gamma analysis with 2%/2 mm passing criteria. To assess and compare the LRA between EBT3 and EBT-XD, 21 × 21 cm{sup 2} open fields delivered doses of 1000, 2000, and 3000 cGy to both types of film. Films were placed at the center of the scanner, and ratios of measured to TPS predicted doses were calculated at 50 and 80 mm lateral from the scanner center in order to quantitatively assess the LRA. To evaluate the energy dependence of EBT-XD film, seven known doses ranging from 400 to 3000 cGy were delivered using both 6 and 18 MV beams and the resulting optical densities (ODs) compared. Results: The gamma passing rate was 99.1% for the 6 MV EDW delivery. EBT-XD film exhibited minimal LRA (<1%) up to 3000 cGy. In contrast, EBT3 demonstrated an under-response of 11.3% and 22.7% at lateral positions of 50 and 80 mm, respectively, for the 3000 cGy exposure. Differences between ODs of the EBT-XD films exposed to known doses from 6 to 18 MV beams were <0.8% suggesting minimal energy dependence throughout this energy range. Conclusions: The LRA of EBT-XD is greatly reduced when compared to EBT3. This in combination with its accuracy from 0 to 3000 cGy and minimal energy dependence from 6 to 18 MV makes EBT-XD film well suited for dosimetric measurements in high dose SRS/SBRT applications.

  6. Technical Note: Initial characterization of the new EBT-XD Gafchromic film

    International Nuclear Information System (INIS)

    Grams, Michael P.; Gustafson, Jon M.; Long, Kenneth M.; Fong de los Santos, Luis E.

    2015-01-01

    Purpose: To assess the dosimetric accuracy and energy dependence of the new EBT-eXtended Dose (XD) Gafchromic film and to compare the lateral response artifact (LRA) between EBT-XD and EBT3 film. Methods: EBT3 and EBT-XD calibration curves were created by exposing films to known doses from 0 to 3000 cGy using a 6 MV beam. To assess the accuracy and dynamic range of EBT-XD, a 60° enhanced dynamic wedge (EDW) was used to deliver a dose range of approximately 200–2900 cGy. Comparison to treatment planning system (TPS) calculation was made using a gamma analysis with 2%/2 mm passing criteria. To assess and compare the LRA between EBT3 and EBT-XD, 21 × 21 cm 2 open fields delivered doses of 1000, 2000, and 3000 cGy to both types of film. Films were placed at the center of the scanner, and ratios of measured to TPS predicted doses were calculated at 50 and 80 mm lateral from the scanner center in order to quantitatively assess the LRA. To evaluate the energy dependence of EBT-XD film, seven known doses ranging from 400 to 3000 cGy were delivered using both 6 and 18 MV beams and the resulting optical densities (ODs) compared. Results: The gamma passing rate was 99.1% for the 6 MV EDW delivery. EBT-XD film exhibited minimal LRA (<1%) up to 3000 cGy. In contrast, EBT3 demonstrated an under-response of 11.3% and 22.7% at lateral positions of 50 and 80 mm, respectively, for the 3000 cGy exposure. Differences between ODs of the EBT-XD films exposed to known doses from 6 to 18 MV beams were <0.8% suggesting minimal energy dependence throughout this energy range. Conclusions: The LRA of EBT-XD is greatly reduced when compared to EBT3. This in combination with its accuracy from 0 to 3000 cGy and minimal energy dependence from 6 to 18 MV makes EBT-XD film well suited for dosimetric measurements in high dose SRS/SBRT applications

  7. Outdoor solar UVA dose assessment with EBT2 radiochromic film using spectrophotometer and densitometer measurements

    International Nuclear Information System (INIS)

    Abukassem, I.; Bero, M.A.

    2015-01-01

    Direct measurements of solar ultraviolet radiations (UVRs) have an important role in the protection of humans against UVR hazard. This work presents simple technique based on the application of EBT2 GAFCHROMIC R film for direct solar UVA dose assessment. It demonstrates the effects of different parts of the solar spectrum (UVB, visible and infrared) on performed UVA field measurements and presents the measurement uncertainty budget. The gradient of sunlight exposure level permitted the authors to establish the mathematical relationships between the measured solar UVA dose and two measured quantities: the first was the changes in spectral absorbance at the wavelength 633 nm (A 633 ) and the second was the optical density (OD). The established standard relations were also applied to calculate the solar UVA dose variations during the whole day; 15 min of exposure each hour between 8:00 and 17:00 was recorded. Results show that both applied experimental methods, spectrophotometer absorbance and densitometer OD, deliver comparable figures for EBT2 solar UVA dose assessment with relative uncertainty of 11 % for spectral absorbance measurements and 15 % for OD measurements. (authors)

  8. Evaluation of Multiple-Sampling Function used with a Microtek flatbed scanner for Radiation Dosimetry Calibration of EBT2 Film

    Energy Technology Data Exchange (ETDEWEB)

    Chang, Liyun [Department of Medical Imaging and Radiological Sciences, I-Shou University, Kaohsiung 82445, Taiwan (China); Ho, Sheng-Yow [Department of Nursing, Chang Jung Christian University, Tainan 71101, Taiwan (China); Department of Radiation Oncology, Chi Mei Medical Center, Liouying, Tainan 73657, Taiwan (China); Ding, Hueisch-Jy [Department of Medical Imaging and Radiological Sciences, I-Shou University, Kaohsiung 82445, Taiwan (China); Hwang, Ing-Ming [Department of Medical Imaging and Radiology, Shu Zen College of Medicine and Management, Kaohsiung 82144, Taiwan (China); Chen, Pang-Yu, E-mail: pangyuchen@yahoo.com.tw [Department of Radiation Oncology, Sinlau Christian Hospital, Tainan 70142, Taiwan (China); Lee, Tsair-Fwu, E-mail: tflee@kuas.edu.tw [Medical Physics and Informatics Laboratory, Department of Electronics Engineering, National Kaohsiung University of Applied Sciences, Kaohsiung 80778, Taiwan (China)

    2016-10-01

    The radiochromic EBT2 film is a widely used quality assurance device for radiation therapy. This study evaluated the film calibration performance of the multiple-sampling function, a function of the ScanWizard Pro scanning software provided by the manufacturer, when used with Microtek 9800XL plus (9800XL{sup +}) flatbed scanner. By using the PDD method, each one of the eight EBT2 films, four delivered by 290 monitor unit (MU) and four by 88 MU via 6-MV photon beams, was tightly sandwiched in a 30{sup 3}-cm{sup 3} water equivalent polystyrene phantom prior to irradiation. Before and after irradiation, all films were scanned using the Microtek 9800XL{sup +} scanner with five different modes of the multiple-sampling function, which could generate the image with the averaged result of multiple-sampling. The net optical densities (netOD) on the beam central axis of film were assigned to corresponding depth doses for calibration. For each sampling mode with either delivered MU, the depth-dose uncertainty of a single film from repeated scans and that of a single scan of the four films were analyzed. Finally, the calibration error and the combined calibration uncertainty between film determined depth-doses and delivered depth-doses were calculated and evaluated for each sampling mode. All standard deviations and the calibration error were demonstrated to be unrelated to the number of sampling lines. The calibration error of the 2-line and 16-line mode was within 3 cGy and better than that of the other modes. The combined uncertainty of the 2-line mode was the lowest, which was generally less than 6 cGy except for the delivered dose around 100 cGy. The evaluation described herein revealed that the EBT2 film calibrated with the 2-line mode has relatively lower error, scanning time and combined uncertianty. Therefore, it is recommended for routine EBT2 film calibration and verification of treatment plans.

  9. Evaluation of Multiple-Sampling Function used with a Microtek flatbed scanner for Radiation Dosimetry Calibration of EBT2 Film

    International Nuclear Information System (INIS)

    Chang, Liyun; Ho, Sheng-Yow; Ding, Hueisch-Jy; Hwang, Ing-Ming; Chen, Pang-Yu; Lee, Tsair-Fwu

    2016-01-01

    The radiochromic EBT2 film is a widely used quality assurance device for radiation therapy. This study evaluated the film calibration performance of the multiple-sampling function, a function of the ScanWizard Pro scanning software provided by the manufacturer, when used with Microtek 9800XL plus (9800XL + ) flatbed scanner. By using the PDD method, each one of the eight EBT2 films, four delivered by 290 monitor unit (MU) and four by 88 MU via 6-MV photon beams, was tightly sandwiched in a 30 3 -cm 3 water equivalent polystyrene phantom prior to irradiation. Before and after irradiation, all films were scanned using the Microtek 9800XL + scanner with five different modes of the multiple-sampling function, which could generate the image with the averaged result of multiple-sampling. The net optical densities (netOD) on the beam central axis of film were assigned to corresponding depth doses for calibration. For each sampling mode with either delivered MU, the depth-dose uncertainty of a single film from repeated scans and that of a single scan of the four films were analyzed. Finally, the calibration error and the combined calibration uncertainty between film determined depth-doses and delivered depth-doses were calculated and evaluated for each sampling mode. All standard deviations and the calibration error were demonstrated to be unrelated to the number of sampling lines. The calibration error of the 2-line and 16-line mode was within 3 cGy and better than that of the other modes. The combined uncertainty of the 2-line mode was the lowest, which was generally less than 6 cGy except for the delivered dose around 100 cGy. The evaluation described herein revealed that the EBT2 film calibrated with the 2-line mode has relatively lower error, scanning time and combined uncertianty. Therefore, it is recommended for routine EBT2 film calibration and verification of treatment plans.

  10. Evaluation of a fluorescent light densitometer for radiochromic film analysis

    Energy Technology Data Exchange (ETDEWEB)

    Cheung Tsang; Butson, M.J. E-mail: butsonm@iahs.nsw.gov.au; Yu, P.K.N

    2002-01-01

    Radiochromic film dosimetry used in high energy X-ray detection has been evaluated using a VIDAR VXR-12 digitiser for accuracy and results are compared to other densitometers. The VIDAR scanner uses a broad band fluorescent light source which however, produces a negligible ultraviolet reaction effect on the radiochromic film with an estimated equivalent dose of 0.25 cGy darkening per scan to MD-55-2 film. By varying the exposure setting on the scanner, more accurate information can be obtained using radiochromic film within a certain optical density range. Over the range of 0-20 Gy, the scanner can accurately measure dose with a standard deviation of 1.8% calculated between the measured values and a polynomial fit to data. The VIDAR scanner is shown to be a suitable densitometer for radiochromic film.

  11. Evaluation of a fluorescent light densitometer for radiochromic film analysis

    International Nuclear Information System (INIS)

    Cheung Tsang; Butson, M.J.; Yu, P.K.N.

    2002-01-01

    Radiochromic film dosimetry used in high energy X-ray detection has been evaluated using a VIDAR VXR-12 digitiser for accuracy and results are compared to other densitometers. The VIDAR scanner uses a broad band fluorescent light source which however, produces a negligible ultraviolet reaction effect on the radiochromic film with an estimated equivalent dose of 0.25 cGy darkening per scan to MD-55-2 film. By varying the exposure setting on the scanner, more accurate information can be obtained using radiochromic film within a certain optical density range. Over the range of 0-20 Gy, the scanner can accurately measure dose with a standard deviation of 1.8% calculated between the measured values and a polynomial fit to data. The VIDAR scanner is shown to be a suitable densitometer for radiochromic film

  12. Study of the Radiochromic Film for High Dose Measurement in Radiation Processing

    Directory of Open Access Journals (Sweden)

    CHEN Yi-zhen

    2016-02-01

    Full Text Available To establish the radiochromic film dosimeter for high dose level measurement during radiation processing, By corresponding formula and its preparation process research, batches of radiochromic film dosimeters were prepared using nylon as substrate and pararosaniline cyanide as dye. In Co-60 gamma reference radiation field, dosimetry response performance of radiochromic film was studied and results showed that the repeatability was good to 1.0%. The response curves demonstrated good linearity in the dose range of 5-210 kGy, and the signal of radiochromic film dosimeters after irradiation under the condition of low temperature storage within 2 weeks was stable. In addition, the radiochromic film dosimeters were not found to have noticeable dose rate dependence in the range of this experiment. In the linear dose range, radiochromic film dosimeter measures the absorbed dose, with extended uncertainty 4.2% (k=2 for Co-60 gamma rays. The film was suitable as dosimeters for the parameters measurement of the electron beam on the accelerator.

  13. Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs

    Energy Technology Data Exchange (ETDEWEB)

    Fiandra, Christian; Fusella, Marco; Filippi, Andrea Riccardo; Ricardi, Umberto; Ragona, Riccardo [Department of Oncology, Radiation Oncology Unit, University of Torino, Turin 10126 (Italy); Giglioli, Francesca Romana [Medical Physics Unit, Azienda Ospedaliera Città della Salute e della Scienza, Turin 10126 (Italy); Mantovani, Cristina [Radiation Oncology Department, Azienda Ospedaliera Città della Salute e della Scienza, Turin 10126 (Italy)

    2013-08-15

    Purpose: Patient-specific quality assurance in volumetric modulated arc therapy (VMAT) brain stereotactic radiosurgery raises specific issues on dosimetric procedures, mainly represented by the small radiation fields associated with the lack of lateral electronic equilibrium, the need of small detectors and the high dose delivered (up to 30 Gy). Gafchromic{sup TM} EBT2 and EBT3 films may be considered the dosimeter of choice, and the authors here provide some additional data about uniformity correction for this new generation of radiochromic films.Methods: A new analysis method using blue channel for marker dye correction was proposed for uniformity correction both for EBT2 and EBT3 films. Symmetry, flatness, and field-width of a reference field were analyzed to provide an evaluation in a high-spatial resolution of the film uniformity for EBT3. Absolute doses were compared with thermoluminescent dosimeters (TLD) as baseline. VMAT plans with multiple noncoplanar arcs were generated with a treatment planning system on a selected pool of eleven patients with cranial lesions and then recalculated on a water-equivalent plastic phantom by Monte Carlo algorithm for patient-specific QA. 2D quantitative dose comparison parameters were calculated, for the computed and measured dose distributions, and tested for statistically significant differences.Results: Sensitometric curves showed a different behavior above dose of 5 Gy for EBT2 and EBT3 films; with the use of inhouse marker-dye correction method, the authors obtained values of 2.5% for flatness, 1.5% of symmetry, and a field width of 4.8 cm for a 5 × 5 cm{sup 2} reference field. Compared with TLD and selecting a 5% dose tolerance, the percentage of points with ICRU index below 1 was 100% for EBT2 and 83% for EBT3. Patients analysis revealed statistically significant differences (p < 0.05) between EBT2 and EBT3 in the percentage of points with gamma values <1 (p= 0.009 and p= 0.016); the percent difference as well as

  14. Evolution, present state and future of the radiochromic dyeing films

    International Nuclear Information System (INIS)

    Villarreal B, J.E.

    2000-01-01

    The evolution of radiochromic films, their present state and their developing future are the object of this work. For this purpose a review of the evolution was realized and also of the present state of using the radiochromic dyeing films as tools to determine the absorbed doses distribution produces by beta emissor sources, beta-gamma, electrons and X-rays. In particular it is presented the development of radiochromic films type GafChromic that by their quality in terms of reproducibility, sensibility and high spatial resolution they have been converted in those films of greatest use so dominating market. Since one of the application fields more important of this type of films is clinical dosimetry, the more eminent applications in this area are presented, which the high resolution dosimetry that use GafChromic has been converted in a basic tool. On the other hand the scopes of this type of dosimetry and the possible development lines of dosimetry with radiochromic dyeing films are discussed. (Author)

  15. Quantitative investigation of the effects of the scanning parameters in the digitization of EBT and EBT2 Gafchromic film dosimetry with flatbed scanners.

    Science.gov (United States)

    Hu, Yida; Ahmad, Salahuddin; Ali, Imad

    2012-01-01

    With increasing popularity and complexity of intensity-modulated radiation therapy (IMRT) delivery modalities including regular and arc therapies, there is a growing challenge for validating the accuracy of dose distributions. Gafchromic films have superior characteristics for dose verification over other conventional dosimeters. In order to optimize the use of Gafchromic films in clinical IMRT quality assurance procedures, the scanning parameters of EBT and EBT2 films with a flatbed scanner were investigated. The effects of several parameters including scanning position, orientation, uniformity, film sensitivity and optical density (OD) growth after irradiation were quantified. The profiles of the EBT and EBT2 films had a noise level of 0.6% and 0.7%, respectively. Considerable orientation dependence was observed and the scanner value difference between landscape and portrait modes were about 12% and 10% for EBT and EBT2 films, respectively. The highest response sensitivity was observed using digitized red color images of the EBT2 film scanned with landscape mode. The total system non-uniformity composed of contributions from the film and the scanner was less than 1.7%. OD variations showed that EBT gray scale grew slower, however, reached higher growth values of 15% when compared with EBT2 gray scale which grew 12% after a long time (480 hours) post-irradiation. The EBT film using the red color channel showed the minimal growth where OD increased up to 3% within 3 days after irradiation, and took one week to stabilize.

  16. Development and implementation of own software for dosimetry multichannel film

    International Nuclear Information System (INIS)

    Jimenez Feltstrom, D.; Reyes Garcia, R.; Luis Simon, F. J.; Carrasco Herrera, M.; Sanchez Carmona, G.; Herrador Cordoba, M.

    2013-01-01

    The objective of this work is to develop its own software for multichannel film dosimetry Radiochromic EBT2. Compare the results obtained with its use in multichannel and single-channel dosimetry. Check that the multi-channel dosimetry eliminates much of the artifacts caused by dirt, fingerprints, scratches, etc. Radiochromic in film and scanner devices. (Author)

  17. SU-E-T-498: Initial Characterization of the New EBT XD Gafchromic Film

    International Nuclear Information System (INIS)

    Grams, M; Gustafson, J; Long, K; Fong de los Santos, L

    2015-01-01

    Purpose To assess the dosimetric accuracy of the new EBT XD (eXtended Dose) Gafchromic film and to compare the lateral scan effect between EBT XD and EBT3 film. Methods: All films were exposed in solid water using a 6 MV beam and were analyzed with FilmQA Pro software and an Epson 10000 XL scanner. Calibration was performed using films exposed to 8 known doses from 0–3000 cGy. To assess dosimetric accuracy, a 15×18 cm 2 60° EDW delivered a dose range of 200–2900 cGy to 4 separate EBT XD films. Comparison to treatment planning system (TPS) predictions was made using a gamma analysis with 2%/2mm passing criteria. To assess the lateral scan effect, 21 × 21 cm 2 open fields delivered doses of 1000, 2000, and 3000 cGy isocentrically to both EBT3 and EBT XD films at 10 cm depth. Films were centered and scanned such that the widest parts of the exposed film extended lateral to the center of the scanner bed in order to maximize the lateral scan effect. Ratios of measured to TPS predicted doses for EBT3 and EBT XD film were calculated at 5 cm and 8 cm lateral from the center of the scanner bed to quantitatively assess the lateral scan effect. Results The average gamma passing rate for the EDW was 98.9% (range 98.7 – 99.1%). EBT XD film exhibited minimal lateral scan effect (<1%) up to 30 Gy. In contrast, EBT3 showed an under-response of 13% and 25% at lateral positions of 5 cm and 8 cm respectively for the 30 Gy exposure. Conclusion EBT XD can accurately measure dose over a wide dynamic range. The non-polarizing structure of its active ingredient essentially eliminates the lateral scan effect up to at least 30 Gy. EBT XD film is therefore well suited for SRS/SBRT verification measurements

  18. SU-E-T-498: Initial Characterization of the New EBT XD Gafchromic Film

    Energy Technology Data Exchange (ETDEWEB)

    Grams, M; Gustafson, J; Long, K; Fong de los Santos, L [Mayo Clinic, Rochester, MN (United States)

    2015-06-15

    Purpose To assess the dosimetric accuracy of the new EBT XD (eXtended Dose) Gafchromic film and to compare the lateral scan effect between EBT XD and EBT3 film. Methods: All films were exposed in solid water using a 6 MV beam and were analyzed with FilmQA Pro software and an Epson 10000 XL scanner. Calibration was performed using films exposed to 8 known doses from 0–3000 cGy. To assess dosimetric accuracy, a 15×18 cm{sup 2} 60° EDW delivered a dose range of 200–2900 cGy to 4 separate EBT XD films. Comparison to treatment planning system (TPS) predictions was made using a gamma analysis with 2%/2mm passing criteria. To assess the lateral scan effect, 21 × 21 cm{sup 2} open fields delivered doses of 1000, 2000, and 3000 cGy isocentrically to both EBT3 and EBT XD films at 10 cm depth. Films were centered and scanned such that the widest parts of the exposed film extended lateral to the center of the scanner bed in order to maximize the lateral scan effect. Ratios of measured to TPS predicted doses for EBT3 and EBT XD film were calculated at 5 cm and 8 cm lateral from the center of the scanner bed to quantitatively assess the lateral scan effect. Results The average gamma passing rate for the EDW was 98.9% (range 98.7 – 99.1%). EBT XD film exhibited minimal lateral scan effect (<1%) up to 30 Gy. In contrast, EBT3 showed an under-response of 13% and 25% at lateral positions of 5 cm and 8 cm respectively for the 30 Gy exposure. Conclusion EBT XD can accurately measure dose over a wide dynamic range. The non-polarizing structure of its active ingredient essentially eliminates the lateral scan effect up to at least 30 Gy. EBT XD film is therefore well suited for SRS/SBRT verification measurements.

  19. Outdoor solar UVA dose assessment with EBT2 radiochromic film using spectrophotometer and densitometer measurements.

    Science.gov (United States)

    Abukassem, I; Bero, M A

    2015-04-01

    Direct measurements of solar ultraviolet radiations (UVRs) have an important role in the protection of humans against UVR hazard. This work presents simple technique based on the application of EBT2 GAFCHROMIC(®) film for direct solar UVA dose assessment. It demonstrates the effects of different parts of the solar spectrum (UVB, visible and infrared) on performed UVA field measurements and presents the measurement uncertainty budget. The gradient of sunlight exposure level permitted the authors to establish the mathematical relationships between the measured solar UVA dose and two measured quantities: the first was the changes in spectral absorbance at the wavelength 633 nm (A633) and the second was the optical density (OD). The established standard relations were also applied to calculate the solar UVA dose variations during the whole day; 15 min of exposure each hour between 8:00 and 17:00 was recorded. Results show that both applied experimental methods, spectrophotometer absorbance and densitometer OD, deliver comparable figures for EBT2 solar UVA dose assessment with relative uncertainty of 11% for spectral absorbance measurements and 15% for OD measurements. © The Author 2014. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  20. Dosimetry using radiochromic film and planning algorithms in heterogeneous phantoms; Estudo dosimetrico com filme radiocromico e algoritmos de planejamento em fantomas heterogeneos

    Energy Technology Data Exchange (ETDEWEB)

    Leite, Vinicius Freitas

    2012-07-01

    This work analyzes, through the study of the interaction of electromagnetic radiation with matter, two schemes of heterogeneous phantoms schematised to simulate real cases of planning with different electronic densities through the Pencil Beam, Collapsed Cone and Analytical Anisotropic Algorithm algorithms and compare with measurements Of relative absorbed dose in an IBA CC13 ionization chamber and Gafchromic® EBT2 radiochromic film. Epichlorohydrin rubber and its compatibility in comparison with human bone has also been evaluated. The assembly of the heterogeneous phantoms was feasible and the results regarding the density and attenuation of the rubber presented consistent values. However, the study of PDPs in constructed phantoms showed a considerable percentage discrepancy between measurements and planning.

  1. Radiochromic dye film for ionizing and non ionizing radiation dosimetry

    International Nuclear Information System (INIS)

    Said, F.I.; Elbahay, A.Z.

    1984-01-01

    During the present study different types of radiochromic dye films have been investigated for γ-ray as well as ultraviolet radiation dosimetry. The technique is based on the change in absorption spectra for dosed samples with special emphasis at optical absorption bands observed at 510nm and 605nm. Dose response of different types of radiochromic dye films was determined in the range between 1 to 40 KGy γ rays. The technique is used for γ-ray dosimetry in processing with high quality. The increasing use of U.V. light in industry and medicine makes it necessary to have simple practical methods for U.V. dosimetry. To this purpose radiochromic dye films are suggested for the first time. The response was investigated for emission consisted mainly of 254 and 366nm photons over the range from 50-700 m W.S. cm -2 . The experimental results indicate that radiochromic dye films are practical and simple technique for γ-ray and U.V. dosimetry for industrial and medical applications

  2. A novel method for alpha dosimetry using peeled-off Gafchromic EBT3 films

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Kwang-Ho; Kim, Eun-Hee [Seoul National University, Seoul (Korea, Republic of)

    2016-10-15

    One can estimate dose imposed to film by measuring the optical density of film. EBT3 film has been used in dose measurement for photon, proton, and electron beams but not for alpha particles. Both sides of diacetylene monomer layer are covered with 100 μm-thick polyester coating layers, through which alpha particles even at several MeV cannot penetrate. A recent study demonstrated the use of EBT3 film in alpha dosimetry by peeling off one side of polyester coating layer. Their study did not inform the reliability of measurement using the peeled-off films. In this study, we evaluated the feasibility of EBT3 film as a substitute for conventional alpha dosimeters and checked the uncertainty of dose measurements obtained with peeled-off EBT3 films. We also applied this film dosimeter to measuring of the fluence distribution at cell targets in a culture dish set in the alpha irradiation chamber of the Radiation Bioengineering Laboratory (RadBio Lab) at Seoul National University (SNU). In this work, we confirmed the feasibility of using Gafchromic EBT3 films for alpha dosimetry. The peeled-off EBT3 films can make a convenient alpha dosimeter by carrying an uncertainty less than 9 %.

  3. Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry.

    Science.gov (United States)

    Desroches, Joannie; Bouchard, Hugo; Lacroix, Frédéric

    2010-04-01

    The purpose of this study is to determine the effect on the measured optical density of scanning on either side of a Gafchromic EBT and EBT2 film using an Epson (Epson Canada Ltd., Toronto, Ontario) 10000XL flat bed scanner. Calibration curves were constructed using EBT2 film scanned in landscape orientation in both reflection and transmission mode on an Epson 10000XL scanner. Calibration curves were also constructed using EBT film. Potential errors due to an optical density difference from scanning the film on either side ("face up" or "face down") were simulated. Scanning the film face up or face down on the scanner bed while keeping the film angular orientation constant affects the measured optical density when scanning in reflection mode. In contrast, no statistically significant effect was seen when scanning in transmission mode. This effect can significantly affect relative and absolute dose measurements. As an application example, the authors demonstrate potential errors of 17.8% by inverting the film scanning side on the gamma index for 3%-3 mm criteria on a head and neck intensity modulated radiotherapy plan, and errors in absolute dose measurements ranging from 10% to 35% between 2 and 5 Gy. Process consistency is the key to obtaining accurate and precise results in Gafchromic film dosimetry. When scanning in reflection mode, care must be taken to place the film consistently on the same side on the scanner bed.

  4. Radiochromic film calibration for dosimetry in computed tomography tests

    International Nuclear Information System (INIS)

    Costa, K. C.; Prata M, A.; Ladino G, A. M.; Costa, K. L.

    2017-10-01

    Radiochromic film applications in dosimetry have become increasingly significant for studies on radiotherapy and diagnostic tests. Due to sensitivity to exposure to ionizing radiation, radiochromic films are commonly used to obtain dose distribution maps. The objective of this study is to obtain the calibration curves of the radiographic film for exposure with X-ray beam in a computerized tomography (CT) scanner to realize measures of typical doses found in radiodiagnosis tests. It was used Gafchromic Xr-AQ2 film, which shows little sensitivity to visible light and a response in the range of 0.1 to 20 Gy for X-ray beam in a tube voltage supply range ranging from 20 kV to 200 kV. In the experiments, a head polymethylmethacrylate (PMMA) phantom, with a cylindrical shape with five openings was used. This phantom was placed in the CT scanner isocenter and radiochromic film strips were placed into two openings. The irradiations were performed in a Toshiba Asteion scanner that allows making acquisitions in helical mode. The central slice of the head phantom was irradiated to obtain the values of air kerma in PMMA measured with a pencil ionization chamber. Thereafter, radiochromic film strips were placed into the central and one peripheral opening and 10 cm long scans of the central region of the phantom were carried out with feed voltage of 120 kV. The strips irradiated with different X-ray tube currents were scanned and processed using the ImageJ software to obtain the intensity values resulting from the absorbed radiation by optical density analysis. The calibration curves were obtained for both region, central and peripheral corresponding to the values of air kerma in PMMA measured with ionization chamber. With the curves in hand, CT experiments with applied beams can use radiochromic films as a dosimetry method and then seek the generation of images with lower dose deposition and higher diagnostic quality. (Author)

  5. Radiochromic film calibration for dosimetry in computed tomography tests

    Energy Technology Data Exchange (ETDEWEB)

    Costa, K. C.; Prata M, A. [Federal Center for Technological Education of Minas Gerais, Biomedical Engineering Center, Av. Amazonas 5253, Nova Suica, 30421-169 Belo Horizonte, Minas Gerais (Brazil); Ladino G, A. M. [Federal University of Minas Gerais, Department of Nuclear Engineering, Av. Antonio Carlos 6627, Pampulha, 31270-90 Belo Horizonte, Minas Gerais (Brazil); Costa, K. L., E-mail: apratabhz@gmail.com [University of Itauna, Medicine Department, Rodovia Mg 431 Km 45 s/n, El Dorado, 35680-142 Itauna, Minas Gerais (Brazil)

    2017-10-15

    Radiochromic film applications in dosimetry have become increasingly significant for studies on radiotherapy and diagnostic tests. Due to sensitivity to exposure to ionizing radiation, radiochromic films are commonly used to obtain dose distribution maps. The objective of this study is to obtain the calibration curves of the radiographic film for exposure with X-ray beam in a computerized tomography (CT) scanner to realize measures of typical doses found in radiodiagnosis tests. It was used Gafchromic Xr-AQ2 film, which shows little sensitivity to visible light and a response in the range of 0.1 to 20 Gy for X-ray beam in a tube voltage supply range ranging from 20 kV to 200 kV. In the experiments, a head polymethylmethacrylate (PMMA) phantom, with a cylindrical shape with five openings was used. This phantom was placed in the CT scanner isocenter and radiochromic film strips were placed into two openings. The irradiations were performed in a Toshiba Asteion scanner that allows making acquisitions in helical mode. The central slice of the head phantom was irradiated to obtain the values of air kerma in PMMA measured with a pencil ionization chamber. Thereafter, radiochromic film strips were placed into the central and one peripheral opening and 10 cm long scans of the central region of the phantom were carried out with feed voltage of 120 kV. The strips irradiated with different X-ray tube currents were scanned and processed using the ImageJ software to obtain the intensity values resulting from the absorbed radiation by optical density analysis. The calibration curves were obtained for both region, central and peripheral corresponding to the values of air kerma in PMMA measured with ionization chamber. With the curves in hand, CT experiments with applied beams can use radiochromic films as a dosimetry method and then seek the generation of images with lower dose deposition and higher diagnostic quality. (Author)

  6. Investigation of the gafchromic film -EBT2: Features for UVR measurements

    International Nuclear Information System (INIS)

    Abukassem, A.; Bero, M.A.

    2015-01-01

    Important improvement in applied ultraviolet radiation (UVR) dosimetry is achieved using passive detector based on chemical or biological products. These kinds of UVR detectors change their optical properties in correlation with the dose. This work aims to investigate the gafchromic film EBT2 properties under high UVA radiation dose comparable with long exposure to solar radiations. Measurements showed that about 90% of UVA radiation beam is absorbed in single films sheet (285 μm thickness). The EBT2 film components show good stability under high ultraviolet radiation dose. The increase in film visible spectrum absorbance, under UVA irradiation, is due to the decrease in the film active and topcoat layers transmittance and not of the polyester layers degradation. The change in film absorbance at 633 nm, after UVA dose of 112 kJ/m 2 , is about two and half times more than the initial value of unexposed film A 0 , 633 nm (A 0 , 633 nm ≅0.6). The phenomenon of post-exposure stability for the studied EBT2 film is found reproducible and has a small impact on the measurement accuracy ( ≅1%). The studied EBT2 film absorbance changes depend on the applied UVR dose and not on the irradiance level. Relative divergence between the film responses, measured at different dose rates, is lower than 5% for a wide irradiance range. This dependency is justified by the variation of local responsivity in the film and also the irradiation source stability. The response linearity of the gafchromic film EBT2 has been confirmed over a wide dose rate range in the UVA spectrum.(author)

  7. SU-E-T-128: Dosimetric Characteristics of Gafchromic EBT3 Films for Megavoltage Photon and Proton Beams.

    Science.gov (United States)

    Hayashi, N; Kato, H; Yada, R; Matsunaga, T; Ehara, I; Adachi, Y

    2012-06-01

    Gafchromic film for quantitative analysis was renewed from EBT2 to EBT3 film in November 2011. The purpose of this study is to investigate the relevant characteristics of EBT3 film for its application in dosimetric verification for IMRT/VMAT or proton therapy. We investigated the characteristics of EBT3 film with comparison of previous EBT2 film. The experiments in this study composed two categories. At first, the photo spectroscopy for the irradiated film was compared between EBT2 and EBT3. The film 1 day after the irradiation was analyzed by a photo spectrometer (SR520: JASCO Corporation, Japan). Secondly, we investigated several calibration curves which obtained by same batch. The films were calibrated by irradiation the films to 13 dose steps. The irradiated films were scanned by a flatbed scanner (ES-10000XL, Epson-Seiko Corporation, Japan). The difference on scan orientation was evaluated alternate portrait and landscape directions. The photon and proton beams were delivered from Clinac 21EX (Varian) and Mitsubishi machine, respectively. The peak absorption wavelength of EBT3 film and its response at all active range were basically same with that of EBT2 film. The peak wavelength of photo absorption in EBT3 was observed at 585 and 634 nm. The fog optical density was increased due to the hazy matte polyester for active layer. However, there is no change the tendency of the calibration curve responding to megavoltage photon and proton beams. The scan orientation dependency of EBT3 film was observed with similar to EBT2 film. The optical density of portrait orientation was 10% higher than that of landscape orientation. The dosimetric characteristics of EBT3 film were basically same with EBT2 film. With regard to the matte polyester, the creation of Newton's rings during scanning procedure was reduced. However, the suitable scan protocol should be used for accurate film dosimetry. © 2012 American Association of Physicists in Medicine.

  8. High dose-rate brachytherapy source position quality assurance using radiochromic film

    International Nuclear Information System (INIS)

    Evans, M.D.C.; Devic, S.; Podgorsak, E.B.

    2007-01-01

    Traditionally, radiographic film has been used to verify high-dose-rate brachytherapy source position accuracy by co-registering autoradiographic and diagnostic images of the associated applicator. Filmless PACS-based clinics that do not have access to radiographic film and wet developers may have trouble performing this quality assurance test in a simple and practical manner. We describe an alternative method for quality assurance using radiochromic-type film. In addition to being easy and practical to use, radiochromic film has some advantages in comparison with traditional radiographic film when used for HDR brachytherapy quality assurance

  9. Evaluation of the magnitude of EBT Gafchromic film polarization effects

    International Nuclear Information System (INIS)

    Cheung, T.; Yu, P.K.N.; Butson, M.J.

    2009-01-01

    Gafchromic EBT film, has become a main dosimetric tools for quantitative evaluation of radiation doses in radiation therapy application. One aspect of variability using EBT Gafchromic film is the magnitude of the orientation effect when analysing the film in landscape or portrait mode. This work has utilized a >99% plane polarized light source a non-polarized diffuse light source to investigate the absolute magnitude of EBT Gafchromic films polarization or orientation effects. Results have shown that using a non-polarized light source produces a negligible orientation effect for EBT Gafchromic film and thus the angle of orientation is not important. However, the film exhibits a significant variation in transmitted optical density with angle of orientation to polarized light producing more than 100% increase, or over a doubling of measured O D for films irradiated with x-rays up to dose levels of 5 Gy. The maximum optical density was found to be in a plane at an angle of 14 0 ± 7 0 (2 S D) when the polarizing sheet is turned clockwise with respect to the film. As the magnitude of the orientation effects follows a sinusoidal shape it becomes more critical for alignment accuracy of the film with respect to the polarizing direction in the anticlockwise direction as this will place the alignment of the polarizing axes on the steeper gradient section of the sinusoidal pattern. An average change of 4.5 % per 5 0 is seen for an anticlockwise polarizer rotation where as the effect is 1.2 % per 5 0 for an clockwise polarizer rotation. This may have consequences to the positional accuracy of placement of the EBT Gafchromic film on a scanner as even a 1 0 alignment error can cause an approximate 1 % error in analysis. The magnitude of the orientation effect is therefore dependant on the degree of polarization of the scanning light source and can range from negligible (diffuse LED light source) through to more than 100% or doubling of O D variation with a fully linear

  10. Evaluation of the magnitude of EBT Gafchromic film polarization effects.

    Science.gov (United States)

    Butson, M J; Cheung, T; Yu, P K N

    2009-03-01

    Gafchromic EBT film, has become a main dosimetric tools for quantitative evaluation of radiation doses in radiation therapy application. One aspect of variability using EBT Gafchromic film is the magnitude of the orientation effect when analysing the film in landscape or portrait mode. This work has utilized a > 99% plane polarized light source and a non-polarized diffuse light source to investigate the absolute magnitude of EBT Gafchromic films polarization or orientation effects. Results have shown that using a non-polarized light source produces a negligible orientation effect for EBT Gafchromic film and thus the angle of orientation is not important. However, the film exhibits a significant variation in transmitted optical density with angle of orientation to polarized light producing more than 100% increase, or over a doubling of measured OD for films irradiated with x-rays up to dose levels of 5 Gy. The maximum optical density was found to be in a plane at an angle of 14 degrees +/- 7 degrees (2 SD) when the polarizing sheet is turned clockwise with respect to the film. As the magnitude of the orientation effect follows a sinusoidal shape it becomes more critical for alignment accuracy of the film with respect to the polarizing direction in the anticlockwise direction as this will place the alignment of the polarizing axes on the steeper gradient section of the sinusoidal pattern. An average change of 4.5% per 5 degrees is seen for an anticlockwise polarizer rotation where as the effect is 1.2% per 5 degrees for an clockwise polarizer rotation. This may have consequences to the positional accuracy of placement of the EBT Gafchromic film on a scanner as even a 1 degree alignment error can cause an approximate 1% error in analysis. The magnitude of the orientation effect is therefore dependant on the degree of polarization of the scanning light source and can range from negligible (diffuse LED light source) through to more than 100% or doubling of OD variation

  11. Experimental comparison of profiles of acquired small fields with ionization chambers, diodes, radiochromic s and TLD films

    International Nuclear Information System (INIS)

    Venencia, D.; Garrigo, E.; Filipuzzi, M.; Germanier, A.

    2014-08-01

    The use of radiation small fields, introduced by new techniques, can bring a considerable uncertainty in the precision of the acquired profiles, due to the conditions of lateral electronic non-equilibrium and the perturbations introduced by the detectors (volume effect and alteration of the charged particles flowing) [Das et al., 2007]. The development of new miniature detectors looks to diminish the uncertainty created by the material and the size of the sensitive volume of the dosimeter. For this reason, comparative measurements for three sizes of square field were carried out (20 mm, 10 mm and 5 mm, of side) using a detectors series: 3 ionization chambers (PTW-31003, IBA-CC04, PTW-31016), 2 diodes (PTW-60012, IBA-Sfd), thermoluminescent detectors micro-cubes of 1 mm of edge (TLD-700) and radiochromic s films EBT-3. These last two were used as reference detectors, due to their spatial high resolution and similar performance with Monte Carlo simulations [Francescon et al., 1998]. So much the thermoluminescent detectors as the radiochromic films resolved the profiles in a similar way. Both diodes responded correctly, but the rest of the detectors overestimated the gloom of the fields, which allows conclude that the used TLD (and both diodes) can resolve field sizes correctly, usually utilized in radio-surgery, without producing significant alterations in the acquired data. (author)

  12. Radiochromic film calibration for the RQT9 quality beam

    Science.gov (United States)

    Costa, K. C.; Gomez, A. M. L.; Alonso, T. C.; Mourao, A. P.

    2017-11-01

    When ionizing radiation interacts with matter it generates energy deposition. Radiation dosimetry is important for medical applications of ionizing radiation due to the increasing demand for diagnostic radiology and radiotherapy. Different dosimetry methods are used and each one has its advantages and disadvantages. The film is a dose measurement method that records the energy deposition by the darkening of its emulsion. Radiochromic films have a little visible light sensitivity and respond better to ionizing radiation exposure. The aim of this study is to obtain the resulting calibration curve by the irradiation of radiochromic film strips, making it possible to relate the darkening of the film with the absorbed dose, in order to measure doses in experiments with X-ray beam of 120 kV, in computed tomography (CT). Film strips of GAFCHROMIC XR-QA2 were exposed according to RQT9 reference radiation, which defines an X-ray beam generated from a voltage of 120 kV. Strips were irradiated in "Laboratório de Calibração de Dosímetros do Centro de Desenvolvimento da Tecnologia Nuclear" (LCD / CDTN) at a dose range of 5-30 mGy, corresponding to the range values commonly used in CT scans. Digital images of the irradiated films were analyzed by using the ImageJ software. The darkening responses on film strips according to the doses were observed and they allowed obtaining the corresponding numeric values to the darkening for each specific dose value. From the numerical values of darkening, a calibration curve was obtained, which correlates the darkening of the film strip with dose values in mGy. The calibration curve equation is a simplified method for obtaining absorbed dose values using digital images of radiochromic films irradiated. With the calibration curve, radiochromic films may be applied on dosimetry in experiments on CT scans using X-ray beam of 120 kV, in order to improve CT acquisition image processes.

  13. New developments in radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Soares, C. G.

    2006-01-01

    NIST has been a pioneer in the use of radiochromic film for medical dosimetry applications. Beginning in 1988 with experiments with 90 Sr/Y ophthalmic applicators, this work has continued into the present. A review of the latest applications is presented, which include high activity low-energy photon source dosimetry and ultra-high resolution film densitometry for dose enhancement near stents and microbeam radiation therapy dosimetry. An exciting recent development is the availability of a new radiochromic emulsion which has been developed for IMRT dosimetry. This emulsion is an order of magnitude more sensitive than was previously available. Measurements of the sensitivity and uniformity of samples of this new film are reported, using a spectrophotometer and two scanning laser densitometers. A unique feature of the new emulsion is that the peak of the absorbance spectrum falls at the wavelength of the HeNe lasers used in the densitometer, maximising sensitivity. When read at a wavelength of 633 nm, sensitivities on the order of 900 mAU Gy -1 were determined for this new film type, compared with about 40 mAU Gy -1 for type HS film, 20 mAU Gy -1 for type MD-55-2 film, and 3 mAU Gy -1 for type HD-810. Film uniformities were found to be good, on the order of 6% peak to peak. However, there is a strong polarisation effect in the samples examined, requiring care in film orientation during readout. (authors)

  14. Evaluation of GAFCHROMIC EBT2 dosimetry for the low dose range using a flat-bed scanner with the reflection mode

    International Nuclear Information System (INIS)

    Gotanda, Tatsuhiro; Katsuda, Toshizo; Akagawa, Takuya; Gotanda, Rumi; Tabuchi, Akihiko; Yamamoto, Kenyu; Kuwano, Tadao; Takedo, Yoshihiro; Yatake, Hidetoshi; Yabunaka, Koichi

    2013-01-01

    Recently developed radiochromic films can easily be used to measure absorbed doses because they do not need development processing and indicate a density change that depends on the absorbed dose. However, in GAFCHROMIC EBT2 dosimetry (GAF-EBT2) as a radiochromic film, the precision of the measurement was compromised, because of non-uniformity problems caused by image acquisition using a flat-bed scanner with a transmission mode. The purpose of this study was to improve the precision of the measurement using a flat-bed scanner with a reflection mode at the low absorbed dose dynamic range of GAF-EBT2. The calibration curves of the absorbed dose versus the film density for GAF-EBT2 were provided. X-rays were exposed in the range between ~0 and 120 mGy in increments of about 12 mGy. The results of the method using a flat-bed scanner with the transmission mode were compared with those of the method using the same scanner with the reflection mode. The results should that the determination coefficients (r 2 ) for the straight-line approximation of the calibration curve using the reflection mode were higher than 0.99, and the gradient using the reflection mode was about twice that of the one using the transmission mode. The non-uniformity error that is produced by a flat-bed scanner with the transmission mode setting could be almost eliminated by converting from the transmission mode to the reflection mode. In light of these findings, the method using a flat-bed scanner with the reflection mode (only using uniform white paper) improved the precision of the measurement for the low absorbed dose range.

  15. Investigation of absorption spectra of Gafchromic EBT2 film's components and their impact on UVR dosimetry

    Science.gov (United States)

    Aydarous, Abdulkadir

    2016-05-01

    The absorption spectra of the EBT2 film's components were investigated in conjunction with its use for UVA dosimetry. The polyester (topside) and adhesive layers of the EBT2 film have been gently removed. Gafchromic™ EBT2 films with and without the protected layers (polyester and adhesive) were exposed to UVR of 365 nm for different durations. Thereafter, the UV-visible spectra were measured using a UV-visible spectrophotometer (Model Spectro Dual Split Beam, UVS-2700). Films were digitized using a Nikon CanoScan 9000F Mark II flatbed scanner. The dosimetric characteristics including film's uniformity, reproducibility and post-irradiation development were investigated. The color development of EBT2 and new modified EBT2 (EBT2-M) films irradiated with UVA was relatively stable (less than 1%) immediately after exposure. Based on this study, the sensitivity of EBT2 to UVR with wavelength between ~350 nm and ~390 nm can significantly be enhanced if the adhesive layer (~25 μm) is removed. The polyester layer plays almost no part on absorbing UVR with wavelength between ~320 nm and ~390 nm. Furthermore, various sensitivities for the EBT2-M film has been established depending on the wavelength of analysis.

  16. SU-E-T-44: Angular Dependence of Surface Dose Enhancement Measured On Several Inhomogeneities Using Radiochromic EBT3 Films

    International Nuclear Information System (INIS)

    Jansen, A; Schoenfeld, A; Poppinga, D; Chofor, N; Poppe, B

    2014-01-01

    Purpose: The quantification of the relative surface dose enhancement in dependence on the angle of incidence and the atomic number Z of the surface material. Methods: Experiments were performed with slabs made of aluminum, titanium, copper, silver, dental gold and lead. The metal slabs with equal sizes of 1.0×8.0×8.8mm 3 were embedded in an Octavius 4D phantom (PTW Freiburg, Germany). Radiochromic EBT3 films were used to measure the surface dose for angles of incidence ranging from 0° to 90°. The setup with the metals slabs at the isocenter was irradiated with acceleration voltages of 6MV and 10MV. Water reference measurements were taken under equal conditions. Results: The surface dose enhancement is highest for angles of incidence below 30° and drops significantly for higher. The surface dose enhancement produced by lead and dental gold at 6MV showed a peak of 65%. At 90°, the surface dose enhancement dropped to 15% for both materials. The surface dose enhancements for silver, copper, titanium and aluminum were 45%, 32%, 22% and 12% at 0°, respectively. At an angle of incidence of 80°, the values dropped to 22%, 18%, 12% und 6%. The values for 10MV were very similar. Lead and dental gold showed peaks of 65% und 60%. Their values dropped to 18% at an angle of 90°. The surface dose enhancements for silver, copper, titanium and aluminum were 45%, 30%, 20% and 8% at 0°. At 80° the values dropped to 30%, 20%, 12% and 5%. A dependence of the magnitude of the surface dose enhancement on the atomic number of the surface material can be seen, which is in consistence with literature. Conclusion: The results show that the surface dose enhancements near implant materials with high Z-values should be taken into consideration in radio therapy, even when the angle of incidence is flat

  17. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Lewis, David; Chan, Maria F.

    2015-01-01

    Purpose: A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA. Methods: EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm 2 pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm 2 sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm 2 open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000XL or 11000XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response(C, D, X) = A L,X + B L,X ⋅Response(L, D, X), where D is dose and the coefficients A L,X and B L,X are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once corrections were applied to compensate for

  18. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry.

    Science.gov (United States)

    Lewis, David; Chan, Maria F

    2015-01-01

    A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA. EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm(2) pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm(2) sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm(2) open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000 XL or 11000 XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response(C, D, X) = A(L,X) + B(L,X) ⋅ Response(L, D, X), where D is dose and the coefficients A(L,X) and B(L,X) are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once corrections were applied to compensate for the

  19. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Lewis, David [RCF Consulting, LLC, 54 Benedict Road, Monroe, Connecticut 06468 (United States); Chan, Maria F., E-mail: chanm@mskcc.org [Memorial Sloan-Kettering Cancer Center, 136 Mountain View Boulevard, Basking Ridge, New Jersey 07920 (United States)

    2015-01-15

    Purpose: A known factor affecting the accuracy of radiochromic film dosimetry is the lateral response artifact (LRA) induced by nonuniform response of a flatbed scanner in the direction perpendicular to the scan direction. This work reports a practical solution to eliminate such artifacts for all forms of dose QA. Methods: EBT3 films from a single production lot (02181401) cut into rectangular 4 × 5 cm{sup 2} pieces, with the long dimension parallel to the long dimension of the original 20.3 × 25.4 cm{sup 2} sheets, were exposed at a depth of 5 cm on a Varian Trilogy at the center of a 20 × 20 cm{sup 2} open field at seven doses between 50 and 1600 cGy using 6 MV photons. These films together with an unexposed film from the same production lot were lined one next to the other on an Epson 10000XL or 11000XL scanner in portrait orientation with their long dimension parallel to the scan direction. Scanned images were then obtained with the line of films positioned at seven discrete lateral locations perpendicular to the scan direction. The process was repeated in landscape orientation and on three other Epson scanners. Data were also collected for three additional production lots of EBT3 film (11051302, 03031401, and 03171403). From measurements at the various lateral positions, the scanner response was determined as a function of the lateral position of the scanned film. For a given color channel X, the response at any lateral position L is related to the response at the center, C, of the scanner by Response(C, D, X) = A{sub L,X} + B{sub L,X}⋅Response(L, D, X), where D is dose and the coefficients A{sub L,X} and B{sub L,X} are determined from the film measurements at the center of the scanner and six other discrete lateral positions. The values at intermediate lateral positions were obtained by linear interpolation. The coefficients were determined for the red, green, and blue color channels, preserving the ability to apply triple-channel dosimetry once

  20. Technical Note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry

    International Nuclear Information System (INIS)

    Desroches, Joannie; Bouchard, Hugo; Lacroix, Frederic

    2010-01-01

    Purpose: The purpose of this study is to determine the effect on the measured optical density of scanning on either side of a Gafchromic EBT and EBT2 film using an Epson (Epson Canada Ltd., Toronto, Ontario) 10000XL flat bed scanner. Methods: Calibration curves were constructed using EBT2 film scanned in landscape orientation in both reflection and transmission mode on an Epson 10000XL scanner. Calibration curves were also constructed using EBT film. Potential errors due to an optical density difference from scanning the film on either side (''face up'' or ''face down'') were simulated. Results: Scanning the film face up or face down on the scanner bed while keeping the film angular orientation constant affects the measured optical density when scanning in reflection mode. In contrast, no statistically significant effect was seen when scanning in transmission mode. This effect can significantly affect relative and absolute dose measurements. As an application example, the authors demonstrate potential errors of 17.8% by inverting the film scanning side on the gamma index for 3%--3 mm criteria on a head and neck intensity modulated radiotherapy plan, and errors in absolute dose measurements ranging from 10% to 35% between 2 and 5 Gy. Conclusions: Process consistency is the key to obtaining accurate and precise results in Gafchromic film dosimetry. When scanning in reflection mode, care must be taken to place the film consistently on the same side on the scanner bed.

  1. Angular dependence on the records of dose in radiochromic films strips

    Energy Technology Data Exchange (ETDEWEB)

    Costa, K. C.; Prata M, A. [Centro Federal de Educacao Tecnologica de Minas Gerais, Centro de Engenharia Biomedica, Av. Amazonas 5253, 30421-169 Nova Suica, Belo Horizonte, Minas Gerais (Brazil); Alonso, T. C. [Centro de Desenvolvimento da Tecnologia Nuclear - CNEN, Av. Pte. Antonio Carlos 6627, 31270-901 Pampulha, Belo Horizonte, Minas Gerais (Brazil); Campo de O, P. M., E-mail: kamilacosta1995@gmail.com [Universidade Federal de Minas Gerais, Departamento de Anatomia e Imagen, Av. Prof. Alfredo Balena 190, 30130-100 Belo Horizonte, Minas Gerais (Brazil)

    2016-10-15

    Radiological images have relevant information both the diagnostic results as to treatment decisions. Then, the diagnostic quality of image that allows a proper analysis should be achieved with the lowest possible deposition of dose in a patient. CT scans produce sectional images that allow the observation of internal structures of the human body without overlap. As in conventional radiology, the contrast which allows obtaining CT images results from the difference in X-ray beam absorption, according to the characteristics of each tissue. The increased of the beam absorption by a tissue means that it appears brighter in the image. In CT scanners, X-ray tube rotates around the patient, and this rotation results in a cross-sectional image of the body. From a sectional image series is possible to obtain a 3-dimensional image that can be viewed from different angles. Among the methods of dose measurement is the use of radiochromic films, which record the energy deposition by darkening its emulsion. The radiochromic films show little sensitivity to visible light and respond better to exposure to ionizing radiation. In this work, strips of the radiochromic film GAFCHROMIC XR-QA2 were irradiated at different angular positions for radiation quality RQT8, defining a beam of X-rays generated from a voltage of 100 kV. The response of radiochromic films depending on the doses was assessed through digital images obtained by H P Scan jet G-4050 scanner. Digital images were analyzed using Image-J software, which allowed obtaining numerical values corresponding to the intensity of darkening for each film. The aim of this study is to evaluate the dose deposition in radiochromic film according to the angular variation in order how is affected the record. So, to examine the use of film strips to record doses in Computed Tomography tests. (Author)

  2. SU-F-T-581: The Effect of Scanning Orientation On New EBT-XD Film Using Vidar and Epson Scanners

    International Nuclear Information System (INIS)

    Chiu-Tsao, S; Chan, M

    2016-01-01

    Purpose: To study the effect of film scanning orientation for new EBT-XD film using Vidar and Epson scanners, compared with EBT3 films. Methods: The EBT-XD (lot#01081501) and EBT3 films (lot#11031501) were cut into 4cm×7cm pieces and each film piece was uniformly exposed to a 6MV photon beam (15×15cm"2) at 5 cm depth and 100cm SAD in polystyrene phantom, one at a time. The doses delivered were 0,2.5,5,10,20,30,40 and 50 Gy for EBT-XD films, and 0,0.5,1,2,4,8,12 and 16 Gy for EBT3 films, respectively. All films were scanned at least 16 hours after irradiation, one at a time, in portrait or landscape orientation, using a Vidar DosimetryPro Advantage Red and an Epson 10000XL scanner (red channel). Each film piece was positioned at the center of a frame (36cm×36cm) with a cutout slightly larger than the film size for scanning. The scanned data (71 dpi), were analyzed using ImageJ and the average in the ROI (3cm×3cm) of each film at the center was obtained. Results: Dose response curves (netOD vs. dose) were established and compared with EBT-XD and EBT3 film in portrait or landscape orientation using these two scanners. The portrait to landscape (P/L) ratio for each film/scanner combination was determined. The EBT-XD film is about three times less sensitive than EBT3 film. The netOD is higher for portrait than landscape orientation for all the film/scanner combinations. For EBT-XD and EBT3, the average P/L ratios are 1.02 and 1.03 using Vidar scanner, and 1.03 and 1.05 using Epson scanner (red channel), respectively. Conclusion: The effect of film scanning orientation for new EBT-XD film using Vidar Advantage Red and Epson 10000XL scanners has been studied, in comparison with EBT3 film. The portrait/landscape ratios are the lowest for EBT-XD film using Vidar Advantage Red scanner among the four film/scanner combinations.

  3. SU-F-T-581: The Effect of Scanning Orientation On New EBT-XD Film Using Vidar and Epson Scanners

    Energy Technology Data Exchange (ETDEWEB)

    Chiu-Tsao, S [Quality MediPhys LLC, Denville, NJ (United States); Chan, M [Memorial Sloan-Kettering Cancer Center, Basking Ridge, NJ (United States)

    2016-06-15

    Purpose: To study the effect of film scanning orientation for new EBT-XD film using Vidar and Epson scanners, compared with EBT3 films. Methods: The EBT-XD (lot#01081501) and EBT3 films (lot#11031501) were cut into 4cm×7cm pieces and each film piece was uniformly exposed to a 6MV photon beam (15×15cm{sup 2}) at 5 cm depth and 100cm SAD in polystyrene phantom, one at a time. The doses delivered were 0,2.5,5,10,20,30,40 and 50 Gy for EBT-XD films, and 0,0.5,1,2,4,8,12 and 16 Gy for EBT3 films, respectively. All films were scanned at least 16 hours after irradiation, one at a time, in portrait or landscape orientation, using a Vidar DosimetryPro Advantage Red and an Epson 10000XL scanner (red channel). Each film piece was positioned at the center of a frame (36cm×36cm) with a cutout slightly larger than the film size for scanning. The scanned data (71 dpi), were analyzed using ImageJ and the average in the ROI (3cm×3cm) of each film at the center was obtained. Results: Dose response curves (netOD vs. dose) were established and compared with EBT-XD and EBT3 film in portrait or landscape orientation using these two scanners. The portrait to landscape (P/L) ratio for each film/scanner combination was determined. The EBT-XD film is about three times less sensitive than EBT3 film. The netOD is higher for portrait than landscape orientation for all the film/scanner combinations. For EBT-XD and EBT3, the average P/L ratios are 1.02 and 1.03 using Vidar scanner, and 1.03 and 1.05 using Epson scanner (red channel), respectively. Conclusion: The effect of film scanning orientation for new EBT-XD film using Vidar Advantage Red and Epson 10000XL scanners has been studied, in comparison with EBT3 film. The portrait/landscape ratios are the lowest for EBT-XD film using Vidar Advantage Red scanner among the four film/scanner combinations.

  4. Development and initial evaluation of a spectral microdensitometer for analysing radiochromic films

    International Nuclear Information System (INIS)

    Lee, K Y; Fung, K L; Kwok, C S

    2004-01-01

    Radiation dose deposited on a radiochromic film is considered as a dose image. A precise image extraction system with commensurate capabilities is required to measure the transmittance of the image and translate it to radiation dose. This paper describes the development of a spectral microdensitometer which has been designed to achieve this goal under the conditions of (a) the linearity and sensitivity of the dose response curve of the radiochromic film being highly dependent on the wavelength of the analysing light, and (b) the inherent high spatial resolution of the film. The microdensitometer consists of a monochromator which provides an analysing light of variable wavelength, a film tray on a high-precision scanning stage, a transmission microscope coupled to a thermoelectrically cooled CCD camera, a microcomputer and corresponding interfaces. The measurement of the transmittance of the radiochromic film is made at the two absorption peaks with maximum sensitivities. The high spatial resolution of the instrument, of the order of micrometres, is achieved through the use of the microscope combined with a measure-and-step technique to cover the whole film. The performance of the instrument in regard to the positional accuracy, system reproducibility and dual-peak film calibration was evaluated. The results show that the instrument fulfils the design objective of providing a precise image extraction system for radiochromic films with micrometre spatial resolution and sensitive dose response

  5. Development and initial evaluation of a spectral microdensitometer for analysing radiochromic films

    Energy Technology Data Exchange (ETDEWEB)

    Lee, K Y [Department of Optometry and Radiography, Hong Kong Polytechnic University, Hong Kong (China); Fung, K L [Department of Optometry and Radiography, Hong Kong Polytechnic University, Hong Kong (China); Kwok, C S [Department of Radioimmunotherapy, City of Hope National Medical Centre, Duarte, CA 91010 (United States)

    2004-11-21

    Radiation dose deposited on a radiochromic film is considered as a dose image. A precise image extraction system with commensurate capabilities is required to measure the transmittance of the image and translate it to radiation dose. This paper describes the development of a spectral microdensitometer which has been designed to achieve this goal under the conditions of (a) the linearity and sensitivity of the dose response curve of the radiochromic film being highly dependent on the wavelength of the analysing light, and (b) the inherent high spatial resolution of the film. The microdensitometer consists of a monochromator which provides an analysing light of variable wavelength, a film tray on a high-precision scanning stage, a transmission microscope coupled to a thermoelectrically cooled CCD camera, a microcomputer and corresponding interfaces. The measurement of the transmittance of the radiochromic film is made at the two absorption peaks with maximum sensitivities. The high spatial resolution of the instrument, of the order of micrometres, is achieved through the use of the microscope combined with a measure-and-step technique to cover the whole film. The performance of the instrument in regard to the positional accuracy, system reproducibility and dual-peak film calibration was evaluated. The results show that the instrument fulfils the design objective of providing a precise image extraction system for radiochromic films with micrometre spatial resolution and sensitive dose response.

  6. SU-E-T-44: A Micro-Raman Spectroscopy Study of the Dose-Dependence of EBT3 GafChromicTM Films for Quantifying the Degree of Molecular Polymerization

    Energy Technology Data Exchange (ETDEWEB)

    Callens, M; Van Den Abeele, K [Department of Physics, Wave Propagation and Signal Processing, KU Leuven KULAK, Kortrijk (Belgium); Crijns, W; Depuydt, T; Maes, F; Haustermans, K [Department of Radiation Oncology, KU Leuven, Leuven (Belgium); Simons, V [IMEC, Leuven (Belgium); De Wolf, I [IMEC, Leuven (Belgium); Department of Materials Engineering, KU Leuven, Leuven (Belgium); D’hooge, J [Department of Cardiovascular Sciences, Bio-Medical Science Group, KU Leuven, Leuven (Belgium); D’Agostino, E [DoseVue NV, Hasselt (Belgium); Pfeiffer, H [Department of Materials Engineering, KU Leuven, Leuven (Belgium)

    2015-06-15

    Purpose: Radiochromic films, such as the poly-diacetylene-based EBT3 GafChromic{sup TM} films (Ashland Specialty Ingredients, Wayne, NY, USA), are widely used for dosimetry applications because of their clear energy independence, high spatial resolution, near tissue equivalence and easy handling. The films undergo a slight color change by radiation-induced polymerization of diacetylene monomers. But more importantly, the film becomes optically less transparent with increasing radiation dose, with a saturation starting between 10 and 20 Gy, i.e. a common SBRT dose level. In contrast to the chromatic properties, less attention has been given to the underlying molecular mechanism that induces this partial color change and strongly reduces the transparency. Therefore, the current work investigates the variation of the molecular composition of the active layer of EBT3 films for an SBRT dose range. Method: Uncoated EBT3 films were irradiated with a 6 MV photon beam using dose levels between 0 and 20 Gy. The relative variation of the polymer content as a function of the applied radiation dose was studied using micro-Raman spectroscopy. Raman spectroscopy with a 633 nm probe laser incident on the active layer allowed to identify the film constituents and to estimate the amount of poly-diacetylenes from the intensities of the unique molecular vibrations of the molecule. Results: The normalized intensity of all polymeric vibrations, and most notably the polymeric triple and double carbon-carbon bonds at 2058 cm{sup −1} and 1446 cm{sup −1} respectively, increase with increasing dose up to a saturation level starting at about 10 Gy, indicating a corresponding increase and saturation of the amount of polymers. This molecular saturation process is the main cause of the non-linear dose response (i.e. a transparency reduction) and of the limited dose range of the studied films. Conclusion: Raman spectroscopy provides new and more fundamental insights in the mechanism of the

  7. Evaluation of two-dimensional bolus effect of immobilization/support devices on skin doses: A radiochromic EBT film dosimetry study in phantom

    International Nuclear Information System (INIS)

    Chiu-Tsao, Sou-Tung; Chan, Maria F.

    2010-01-01

    Purpose: In this study, the authors have quantified the two-dimensional (2D) perspective of skin dose increase using EBT film dosimetry in phantom in the presence of patient immobilization devices during conventional and IMRT treatments. Methods: For 6 MV conventional photon field, the authors evaluated and quantified the 2D bolus effect on skin doses for six different common patient immobilization/support devices, including carbon fiber grid with Mylar sheet, Orfit carbon fiber base plate, balsa wood board, Styrofoam, perforated AquaPlast sheet, and alpha-cradle. For 6 and 15 MV IMRT fields, a stack of two film layers positioned above a solid phantom was exposed at the air interface or in the presence of a patient alpha-cradle. All the films were scanned and the pixel values were converted to doses based on an established calibration curve. The authors determined the 2D skin dose distributions, isodose curves, and cross-sectional profiles at the surface layers with or without the immobilization/support device. The authors also generated and compared the dose area histograms (DAHs) and dose area products from the 2D skin dose distributions. Results: In contrast with 20% relative dose [(RD) dose relative to d max on central axis] at 0.0153 cm in the film layer for 6 MV 10x10 cm 2 open field, the average RDs at the same depth in the film layer were 71%, 69%, 55%, and 57% for Orfit, balsa wood, Styrofoam, and alpha-cradle, respectively. At the same depth, the RDs were 54% under a strut and 26% between neighboring struts of a carbon fiber grid with Mylar sheet, and between 34% and 56% for stretched perforated AquaPlast sheet. In the presence of the alpha-cradle for the 6 MV (15 MV) IMRT fields, the hot spot doses at the effective measurement depths of 0.0153 and 0.0459 cm were 140% and 150% (83% and 89%), respectively, of the isocenter dose. The enhancement factor was defined as the ratio of a given DAH parameter (minimum dose received in a given area) with and without

  8. Characterization of an extrapolation chamber and radiochromic films for verifying the metrological coherence among beta radiation fields; Caracterizacao de uma camara de extrapolacao e filmes radiocromicos para verificacao da coerencia metrologica entre campos padroes de radiacao beta

    Energy Technology Data Exchange (ETDEWEB)

    Castillo, Jhonny Antonio Benavente

    2011-07-01

    The metrological coherence among standard systems is a requirement for assuring the reliability of dosimetric quantities measurements in ionizing radiation field. Scientific and technologic improvements happened in beta radiation metrology with the installment of the new beta secondary standard BSS2 in Brazil and with the adoption of the internationally recommended beta reference radiations. The Dosimeter Calibration Laboratory of the Development Center for Nuclear Technology (LCD/CDTN), in Belo Horizonte, implemented the BSS2 and methodologies are investigated for characterizing the beta radiation fields by determining the field homogeneity, the accuracy and uncertainties in the absorbed dose in air measurements. In this work, a methodology to be used for verifying the metrological coherence among beta radiation fields in standard systems was investigated; an extrapolation chamber and radiochromic films were used and measurements were done in terms of absorbed dose in air. The reliability of both the extrapolation chamber and the radiochromic film was confirmed and their calibrations were done in the LCD/CDTN in {sup 90}Sr/{sup 90}Y, {sup 85}Kr and {sup 147}Pm beta radiation fields. The angular coefficients of the extrapolation curves were determined with the chamber; the field mapping and homogeneity were obtained from dose profiles and isodose with the radiochromic films. A preliminary comparison between the LCD/CDTN and the Instrument Calibration Laboratory of the Nuclear and Energy Research Institute / Sao Paulo (LCI/IPEN) was carried out. Results with the extrapolation chamber measurements showed in terms of absorbed dose in air rates showed differences between both laboratories up to de -I % e 3%, for {sup 90}Sr/{sup 90}Y, {sup 85}Kr and {sup 147}Pm beta radiation fields, respectively. Results with the EBT radiochromic films for 0.1, 0.3 and 0.15 Gy absorbed dose in air, for the same beta radiation fields, showed differences up to 3%, -9% and -53%. The beta

  9. SU-E-T-123: Dosimetric Comparison Between Portrait and Landscape Orientations in Radiochromic Film Dosimetry

    International Nuclear Information System (INIS)

    Kakinohana, Y; Toita, T; Kasuya, G; Ariga, T; Heianna, J; Murayama, S

    2014-01-01

    Purpose: To compare the dosimetric properties of radiochromic films with different orientation. Methods: A sheet of EBT3 film was cut into eight pieces with the following sizes: 15×15 cm2 (one piece), 5x15 cm 2 (two) and 4×5 cm 2 (five). A set of two EBT3 sheets was used at each dose level. Two sets were used changing the delivered doses (1 and 2 Gy). The 5×15 cm 2 pieces were rotated by 90 degrees in relation to each other, such that one had landscape orientation and the other had portrait orientation. All 5×15 cm2 pieces were irradiated with their long side aligned with the x-axis of the radiation field. The 15×15 cm 2 pieces were irradiated rotated at 90 degrees to each other. Five pieces, (a total of ten from two sheets) were used to obtain a calibration curve. The irradiated films were scanned using an Epson ES-2200 scanner and were analyzed using ImageJ software. In this study, no correction was applied for the nonuniform scanner signal that is evident in the direction of the scanner lamp. Each film piece was scanned both in portrait and landscape orientations. Dosimetric comparisons of the beam profiles were made in terms of the film orientations (portrait and landscape) and scanner bed directions (perpendicular and parallel to the scanner movement). Results: In general, portrait orientation exhibited higher noise than landscape and was adversely affected to a great extent by the nonuniformity in the direction of the scanner lamp. A significant difference in the measured field widths between the perpendicular and parallel directions was found for both orientations. Conclusion: Without correction for the nonuniform scanner signal in the direction of the scanner lamp, a landscape orientation is preferable. A more detailed investigation is planned to evaluate quantitatively the effect of orientation on the dosimetric properties of a film

  10. SU-E-T-123: Dosimetric Comparison Between Portrait and Landscape Orientations in Radiochromic Film Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Kakinohana, Y [University of the Ryukyus, Okinawa (Japan); Toita, T; Kasuya, G; Ariga, T; Heianna, J; Murayama, S [University of the Ryukyus, Nishihara-cho, Okinawa (Japan)

    2014-06-01

    Purpose: To compare the dosimetric properties of radiochromic films with different orientation. Methods: A sheet of EBT3 film was cut into eight pieces with the following sizes: 15×15 cm2 (one piece), 5x15 cm{sup 2} (two) and 4×5 cm{sup 2} (five). A set of two EBT3 sheets was used at each dose level. Two sets were used changing the delivered doses (1 and 2 Gy). The 5×15 cm{sup 2} pieces were rotated by 90 degrees in relation to each other, such that one had landscape orientation and the other had portrait orientation. All 5×15 cm2 pieces were irradiated with their long side aligned with the x-axis of the radiation field. The 15×15 cm{sup 2} pieces were irradiated rotated at 90 degrees to each other. Five pieces, (a total of ten from two sheets) were used to obtain a calibration curve. The irradiated films were scanned using an Epson ES-2200 scanner and were analyzed using ImageJ software. In this study, no correction was applied for the nonuniform scanner signal that is evident in the direction of the scanner lamp. Each film piece was scanned both in portrait and landscape orientations. Dosimetric comparisons of the beam profiles were made in terms of the film orientations (portrait and landscape) and scanner bed directions (perpendicular and parallel to the scanner movement). Results: In general, portrait orientation exhibited higher noise than landscape and was adversely affected to a great extent by the nonuniformity in the direction of the scanner lamp. A significant difference in the measured field widths between the perpendicular and parallel directions was found for both orientations. Conclusion: Without correction for the nonuniform scanner signal in the direction of the scanner lamp, a landscape orientation is preferable. A more detailed investigation is planned to evaluate quantitatively the effect of orientation on the dosimetric properties of a film.

  11. Investigation of absorption spectra of Gafchromic EBT2 film's components and their impact on UVR dosimetry

    International Nuclear Information System (INIS)

    Aydarous, Abdulkadir

    2016-01-01

    The absorption spectra of the EBT2 film's components were investigated in conjunction with its use for UVA dosimetry. The polyester (topside) and adhesive layers of the EBT2 film have been gently removed. Gafchromic™ EBT2 films with and without the protected layers (polyester and adhesive) were exposed to UVR of 365 nm for different durations. Thereafter, the UV–visible spectra were measured using a UV–visible spectrophotometer (Model Spectro Dual Split Beam, UVS-2700). Films were digitized using a Nikon CanoScan 9000F Mark II flatbed scanner. The dosimetric characteristics including film's uniformity, reproducibility and post-irradiation development were investigated. The color development of EBT2 and new modified EBT2 (EBT2-M) films irradiated with UVA was relatively stable (less than 1%) immediately after exposure. Based on this study, the sensitivity of EBT2 to UVR with wavelength between ~350 nm and ~390 nm can significantly be enhanced if the adhesive layer (~25 µm) is removed. The polyester layer plays almost no part on absorbing UVR with wavelength between ~320 nm and ~390 nm. Furthermore, various sensitivities for the EBT2-M film has been established depending on the wavelength of analysis. - Highlights: • The absorption spectra of the EBT2 film's components were investigated. • The EBT2 sensitivity to UVR can be enhanced if the adhesive layer is removed. • The polyester layer plays almost no part on absorbing UVA radiation. • The color development of EBT2-M films was almost stable immediately after exposure.

  12. SU-E-T-386: Evaluation of EBT3 Film Response in Different Batches

    Energy Technology Data Exchange (ETDEWEB)

    Escarcia, F [Universidad Nacional Autonoma de Mexico, Mexico Df, DF (Mexico); Herrera, J; Garcia, O [Instituto Nacional de Neurologia y Neurocirugia, Mexico Df, DF (Mexico)

    2015-06-15

    Purpose: To investigate the uniformity of film response of EBT3 film of two different film batches. It has been reported that the response of the EBT film family is not homogenous between film batches. The later may have an impact in the dosimetry of small radiotherapy beams. Methods: A solid water phantom was used for dosimetric measurements. EBT3 film irradiation was performed with a 6 MV photon beam at 5 cm depth with a SAD of 100 cm. All irradiations were performed perpendicularly to the film plane covering the dose range 1 to 10 Gy. Three square field sizes were used to analyze the film response energy dependence: 10, 5 and 1 cm{sup 2}. Two batches of film EBT3 were used #A03181302 (B1) and #03031403 (B2). Film read out was carrying out with an Epson Perfection V750-Pro flatbed scanner in transmission mode with a spatial resolution of 72 dpi, with all post-processing and colour management options turned off, using 48 bits RGB colour depth. The scans were analyzed with the red channel. Results: The results shown that there were differences between the film response for each batch. The differences between batches for 1 Gy were 2%, 6% and 12% for 10, 5 and 1 cm2 square field sizes, respectively. The differences found for 10 Gy were 13%, 14% and 13% for 10, 5 and 1 cm{sup 2} square field sizes, respectively. It can be observed that the dependence with field size dismissed for higher doses. The later may be due to film response saturation at 10 Gy. Conclusion: The EBT3 film -as its predecessors-, it suffer for inter-batch variability in the film response. Further research is required to assess the possible impact in small beam dosimetry.

  13. SU-E-T-386: Evaluation of EBT3 Film Response in Different Batches

    International Nuclear Information System (INIS)

    Escarcia, F; Herrera, J; Garcia, O

    2015-01-01

    Purpose: To investigate the uniformity of film response of EBT3 film of two different film batches. It has been reported that the response of the EBT film family is not homogenous between film batches. The later may have an impact in the dosimetry of small radiotherapy beams. Methods: A solid water phantom was used for dosimetric measurements. EBT3 film irradiation was performed with a 6 MV photon beam at 5 cm depth with a SAD of 100 cm. All irradiations were performed perpendicularly to the film plane covering the dose range 1 to 10 Gy. Three square field sizes were used to analyze the film response energy dependence: 10, 5 and 1 cm 2 . Two batches of film EBT3 were used #A03181302 (B1) and #03031403 (B2). Film read out was carrying out with an Epson Perfection V750-Pro flatbed scanner in transmission mode with a spatial resolution of 72 dpi, with all post-processing and colour management options turned off, using 48 bits RGB colour depth. The scans were analyzed with the red channel. Results: The results shown that there were differences between the film response for each batch. The differences between batches for 1 Gy were 2%, 6% and 12% for 10, 5 and 1 cm2 square field sizes, respectively. The differences found for 10 Gy were 13%, 14% and 13% for 10, 5 and 1 cm 2 square field sizes, respectively. It can be observed that the dependence with field size dismissed for higher doses. The later may be due to film response saturation at 10 Gy. Conclusion: The EBT3 film -as its predecessors-, it suffer for inter-batch variability in the film response. Further research is required to assess the possible impact in small beam dosimetry

  14. Use of radiochromic film as a high-spatial resolution dosimeter by Raman spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Mirza, Jamal Ahmad; Park, Hyeonsuk [Program in Biomedical Radiation Sciences, Department of Transdisciplinary Studies, Graduate School of Convergence Science and Technology, Seoul National University, Seoul 08826 (Korea, Republic of); Park, So-Yeon [Interdisciplinary Program in Radiation Applied Life Sciences, Seoul National University College of Medicine, Seoul 03080 (Korea, Republic of); Ye, Sung-Joon, E-mail: sye@snu.ac.kr [Program in Biomedical Radiation Sciences, Department of Transdisciplinary Studies, Graduate School of Convergence Science and Technology, Seoul National University, Seoul 08826 (Korea, Republic of)

    2016-08-15

    Purpose: Due to increasing demand for high-spatial resolution dosimetry, radiochromic films have been investigated as potential candidates but are often limited by the scanning system, e.g., flatbed optical scanner. In this study, Raman spectroscopy in conjunction with a microscope was selected as an alternative method for high-spatial resolution dosimetry of radiochromic film. Methods: Unlaminated Gafchromic™ EBT3 films were irradiated with doses between 0 and 50 Gy using 6 MV x-rays of a clinical linear accelerator. Depth profiling from the surface of unlaminated film was performed to acquire the maximum Raman intensity peaks of C≡C and C=C stretching bands of diacetylene polymer. The Raman mapping technique for a region of interest (200 × 200, 30 × 30 μm{sup 2}) was developed to reduce a large variation in a Raman spectrum produced with a sampling resolution of a few μm. The preprocessing of Raman spectra was carried out to determine a dosimetric relationship with the amount of diacetylene polymerization. Results: Due to partial diacetylene polymerization upon irradiation, two Raman peaks of C=C and C≡C stretching bands were observed around 1447 and 2060 cm{sup −1}, respectively. The maximum intensities of the two peaks were obtained by positioning a focused laser spot on the surface of unlaminated film. For the dose range of 0–50 Gy, the band heights of both C≡C and C=C peaks increase asymptotically with increasing doses and can be fit with an exponential function of two components. The relative standard deviation in Raman mapping was found to be less than ±5%. By using this technique, dose uniformity was found to be within ±2%. Conclusions: The Raman intensity for C=C and C≡C peaks increases with an increase in the amount of diacetylene polymerization due to an increase in dose. This study shows the potential of Raman spectroscopy as an alternative for absolute dosimetry verifications with a high-spatial resolution of a few μm, but these

  15. A prototype, glassless densitometer traceable to primary optical standards for quantitative radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Rosen, B. S.; Hammer, C. G.; Kunugi, K. A.; DeWerd, L. A.; Soares, C. G.

    2015-01-01

    Purpose: To evaluate a prototype densitometer traceable to primary optical standards and compare its performance to an EPSON Expression ® 10000XL flatbed scanner (the Epson) for quantitative radiochromic film (RCF) dosimetry. Methods: A prototype traceable laser densitometry system (LDS) was developed to mitigate common film scanning artifacts, such as positional scan dependence and high noise in low-dose regions, by performing point-based measurements of RCF suspended in free-space using coherent light. The LDS and the Epson optical absorbance scales were calibrated up to 3 AU, using reference materials calibrated at a primary standards laboratory and a scanner calibration factor (SCF). Calibrated optical density (OD) was determined for 96 Gafchromic ® EBT3 film segments before and after irradiation to one of 16 dose levels between 0 and 10 Gy, exposed to 60 Co in a polymethyl-methacrylate (PMMA) phantom. The sensitivity was determined at each dose level and at two rotationally orthogonal readout orientations to obtain the sensitometric response of each RCF dosimetry system. LDS rotational scanning dependence was measured at nine angles between 0°and 180°, due to the expected interference between coherent light and polarizing EBT3 material. The response curves were fit to the analytic functions predicted by two physical response models: the two-parameter single-hit model and the four-parameter percolation model. Results: The LDS and the Epson absorbance measurements were linear to primary optical standards to within 0.2% and 0.3% up to 2 and 1 AU, respectively. At higher densities, the LDS had an over-response (2.5% at 3 AU) and the Epson an under-response (3.1% and 9.8% at 2 and 3 AU, respectively). The LDS and the Epson SCF over the applicable range were 0.968% ± 0.2% and 1.561% ± 0.3%, respectively. The positional scan dependence was evaluated on each digitizer and shown to be mitigated on the LDS, as compared to the Epson. Maximum EBT3 rotational

  16. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification.

    Science.gov (United States)

    Sankar, A; Ayyangar, Komanduri M; Nehru, R Mothilal; Kurup, P G Gopalakrishna; Murali, V; Enke, Charles A; Velmurugan, J

    2006-01-01

    The quantitative dose validation of intensity-modulated radiation therapy (IMRT) plans require 2-dimensional (2D) high-resolution dosimetry systems with uniform response over its sensitive region. The present work deals with clinical use of commercially available self-developing Radio Chromic Film, Gafchromic EBT film, for IMRT dose verification. Dose response curves were generated for the films using a VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak extended dose range 2 (EDR2) films. The EBT film had a linear response between the dose range of 0 to 600 cGy. The dose-related characteristics of the EBT film, such as post irradiation color growth with time, film uniformity, and effect of scanning orientation, were studied. There was up to 8.6% increase in the color density between 2 to 40 hours after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative differences between calculated and measured dose distributions were analyzed using DTA and Gamma index with the tolerance of 3% dose difference and 3-mm distance agreement. The EDR2 films showed consistent results with the calculated dose distributions, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large-field IMRT verification. For IMRT of smaller field sizes (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films.

  17. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification

    International Nuclear Information System (INIS)

    Sankar, A.; Ayyangar, Komanduri M.; Nehru, R. Mothilal; Gopalakrishna Kurup, P.G.; Murali, V.; Enke, Charles A.; Velmurugan, J.

    2006-01-01

    The quantitative dose validation of intensity-modulated radiation therapy (IMRT) plans require 2-dimensional (2D) high-resolution dosimetry systems with uniform response over its sensitive region. The present work deals with clinical use of commercially available self-developing Radio Chromic Film, Gafchromic EBT film, for IMRT dose verification. Dose response curves were generated for the films using a VXR-16 film scanner. The results obtained with EBT films were compared with the results of Kodak extended dose range 2 (EDR2) films. The EBT film had a linear response between the dose range of 0 to 600 cGy. The dose-related characteristics of the EBT film, such as post irradiation color growth with time, film uniformity, and effect of scanning orientation, were studied. There was up to 8.6% increase in the color density between 2 to 40 hours after irradiation. There was a considerable variation, up to 8.5%, in the film uniformity over its sensitive region. The quantitative differences between calculated and measured dose distributions were analyzed using DTA and Gamma index with the tolerance of 3% dose difference and 3-mm distance agreement. The EDR2 films showed consistent results with the calculated dose distributions, whereas the results obtained using EBT were inconsistent. The variation in the film uniformity limits the use of EBT film for conventional large-field IMRT verification. For IMRT of smaller field sizes (4.5 x 4.5 cm), the results obtained with EBT were comparable with results of EDR2 films

  18. Monte Carlo uncertainty analysis of dose estimates in radiochromic film dosimetry with single-channel and multichannel algorithms.

    Science.gov (United States)

    Vera-Sánchez, Juan Antonio; Ruiz-Morales, Carmen; González-López, Antonio

    2018-03-01

    To provide a multi-stage model to calculate uncertainty in radiochromic film dosimetry with Monte-Carlo techniques. This new approach is applied to single-channel and multichannel algorithms. Two lots of Gafchromic EBT3 are exposed in two different Varian linacs. They are read with an EPSON V800 flatbed scanner. The Monte-Carlo techniques in uncertainty analysis provide a numerical representation of the probability density functions of the output magnitudes. From this numerical representation, traditional parameters of uncertainty analysis as the standard deviations and bias are calculated. Moreover, these numerical representations are used to investigate the shape of the probability density functions of the output magnitudes. Also, another calibration film is read in four EPSON scanners (two V800 and two 10000XL) and the uncertainty analysis is carried out with the four images. The dose estimates of single-channel and multichannel algorithms show a Gaussian behavior and low bias. The multichannel algorithms lead to less uncertainty in the final dose estimates when the EPSON V800 is employed as reading device. In the case of the EPSON 10000XL, the single-channel algorithms provide less uncertainty in the dose estimates for doses higher than four Gy. A multi-stage model has been presented. With the aid of this model and the use of the Monte-Carlo techniques, the uncertainty of dose estimates for single-channel and multichannel algorithms are estimated. The application of the model together with Monte-Carlo techniques leads to a complete characterization of the uncertainties in radiochromic film dosimetry. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  19. Radiochromic film calibration for low-energy seed brachytherapy dose measurement

    Energy Technology Data Exchange (ETDEWEB)

    Morrison, Hali, E-mail: hamorris@ualberta.ca; Menon, Geetha; Sloboda, Ron S. [Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta T6G 1Z2, Canada and Department of Oncology, University of Alberta, Edmonton, Alberta T6G 2R3 (Canada)

    2014-07-15

    Purpose: Radiochromic film dosimetry is typically performed for high energy photons and moderate doses characterizing external beam radiotherapy (XRT). The purpose of this study was to investigate the accuracy of previously established film calibration procedures used in XRT when applied to low-energy, seed-based brachytherapy at higher doses, and to determine necessary modifications to achieve similar accuracy in absolute dose measurements. Methods: Gafchromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 75 kVp, 200 kVp, 6 MV, and (∼28 keV) I-125 photon sources. For the latter irradiations a custom phantom was built to hold a single I-125 seed. Film pieces were scanned with an Epson 10000XL flatbed scanner and the resulting 48-bit RGB TIFF images were analyzed using both FilmQA Pro software andMATLAB. Calibration curves relating dose and optical density via a rational functional form for all three color channels at each irradiation energy were determined with and without the inclusion of uncertainties in the measured optical densities and dose values. The accuracy of calibration curve variations obtained using piecewise fitting, a reduced film measurement area for I-125 irradiation, and a reduced number of dose levels was also investigated. The energy dependence of the film lot used was also analyzed by calculating normalized optical density values. Results: Slight differences were found in the resulting calibration curves for the various fitting methods used. The accuracy of the calibration curves was found to improve at low doses and worsen at high doses when including uncertainties in optical densities and doses, which may better represent the variability that could be seen in film optical density measurements. When exposing the films to doses > 8 Gy, two-segment piecewise fitting was found to be necessary to achieve similar accuracies in absolute dose measurements as when using smaller dose ranges. When reducing the film measurement

  20. An experimental extrapolation technique using the Gafchromic EBT3 film for relative output factor measurements in small x-ray fields

    Energy Technology Data Exchange (ETDEWEB)

    Morales, Johnny E., E-mail: johnny.morales@lh.org.au [Department of Radiation Oncology, Chris O’Brien Lifehouse, 119-143 Missenden Road, Camperdown, NSW 2050, Australia and School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology, Level 4 O Block, Garden’s Point, QLD 4001 (Australia); Butson, Martin; Hill, Robin [Department of Radiation Oncology, Chris O’Brien Lifehouse, 119-143 Missenden Road, Camperdown, NSW 2050, Australia and Institute of Medical Physics, University of Sydney, NSW 2006 (Australia); Crowe, Scott B. [School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology, Level 4 O Block, Garden’s Point, QLD 4001, Australia and Cancer Care Services, Royal Brisbane and Women’s Hospital, Butterfield Street, Herston, QLD 4029 (Australia); Trapp, J. V. [School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology, Level 4 O Block, Garden’s Point, QLD 4001 (Australia)

    2016-08-15

    Purpose: An experimental extrapolation technique is presented, which can be used to determine the relative output factors for very small x-ray fields using the Gafchromic EBT3 film. Methods: Relative output factors were measured for the Brainlab SRS cones ranging in diameters from 4 to 30 mm{sup 2} on a Novalis Trilogy linear accelerator with 6 MV SRS x-rays. The relative output factor was determined from an experimental reducing circular region of interest (ROI) extrapolation technique developed to remove the effects of volume averaging. This was achieved by scanning the EBT3 film measurements with a high scanning resolution of 1200 dpi. From the high resolution scans, the size of the circular regions of interest was varied to produce a plot of relative output factors versus area of analysis. The plot was then extrapolated to zero to determine the relative output factor corresponding to zero volume. Results: Results have shown that for a 4 mm field size, the extrapolated relative output factor was measured as a value of 0.651 ± 0.018 as compared to 0.639 ± 0.019 and 0.633 ± 0.021 for 0.5 and 1.0 mm diameter of analysis values, respectively. This showed a change in the relative output factors of 1.8% and 2.8% at these comparative regions of interest sizes. In comparison, the 25 mm cone had negligible differences in the measured output factor between zero extrapolation, 0.5 and 1.0 mm diameter ROIs, respectively. Conclusions: This work shows that for very small fields such as 4.0 mm cone sizes, a measureable difference can be seen in the relative output factor based on the circular ROI and the size of the area of analysis using radiochromic film dosimetry. The authors recommend to scan the Gafchromic EBT3 film at a resolution of 1200 dpi for cone sizes less than 7.5 mm and to utilize an extrapolation technique for the output factor measurements of very small field dosimetry.

  1. A new method of readout in radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Kellermann, Peer Oliver; Gornik, Erich; Ertl, Adolf

    1998-01-01

    Radiochromic film as a dosimetry medium offers several advantages in high-resolution radiography. A new technique of readout was developed to measure the optical density distributions of the film in purely directed light. This technique implements radiochromic film dosimetry near the film's absorption maximum by using a single-mode top-surface emitting laser diode (675.2 nm). The effective sensitivity of the film, compared with a helium-neon laser densitometer (632.8 nm), is increased approximately threefold. Good accuracy, high spatial resolution and simple assembly of the readout system is achieved. Beam profiles of the four final collimator helmets of a Leksell Gamma Knife (Elekta Inc., Sweden) were experimentally determined. Measured profiles and full-widths at half maximum are consistent with the computer generated data of the dose planning system (Kula 4.4, Elekta Inc., Sweden). The output factor of the 4 mm collimator (the smallest collimator with the steepest dose gradient), essential for the application of well defined doses, was checked. The measurements established an output factor of 826±9 that lies 9±1% lower than the adjusted one. (author)

  2. A new radiochromic dosimeter film

    Science.gov (United States)

    Sidney, L. N.; Lynch, D. C.; Willet, P. S.

    By employing acid-sensitive leuco dyes in a chlorine-containing polymer matrix, a new radiochromic dosimeter film has been developed for gamma, electron beam, and ultraviolet radiation. These dosimeter films undergo a color change from colorless to royal blue, red fuchsia, or black, depending on dye selection, and have been characterized using a visible spectrophotometer over an absorbed dose range of 1 to 100 kGy. The primary features of the film are improved color stability before and after irradiation, whether stored in the dark or under artificial lights, and improved moisture resistance. The effects of absorbed dose, dose rate, and storage conditions on dosimeter performance are discussed. The dosimeter material may be produced as a free film or coated onto a transparent substrate and optionally backed with adhesive. Potential applications for these materials include gamma sterilization indicator films for food and medical products, electron beam dosimeters, and in-line radiation monitors for electron beam and ultraviolet processing.

  3. Sensitivity of film measured off-axis ratios to film calibration curve using radiochromic film

    International Nuclear Information System (INIS)

    Garcia-Hernandez, Diana; Larraga-Gutierrez, Jose M.

    2011-01-01

    Off-axis ratios of conical beams generated with a stereotactic radiosurgery-dedicated Linac were measured with EBT2 film and stereotactic diode. The sensitivity of both full width at half maximum (FWHM) and penumbras (80-20% and 90-10%, respectively), with respect to the characteristics of the film calibration curve fit, was investigated. In all cases, penumbras resulted to be more sensitive than FWHM. However, these differences were, in general, smaller than the ones found between EBT2 reference values and the stereotactic diode measurements. The larger variation in OAR parameters was found to depend on whether the fit intersected or not the origin. A 1D gamma-index analysis showed this difference can be important in all measured conical beams. (author)

  4. Dosimetric properties of Gafchromic (registered) EBT films in medical carbon ion beams

    International Nuclear Information System (INIS)

    MartisIkova, Maria; Jaekel, Oliver

    2010-01-01

    High spatial resolution is desired for dosimetrical verification of patient plans for radiotherapy treatments employing scanned ion beams. This is provided by Gafchromic EBT film, the ancestor of currently available EBT2 films. In this contribution, dosimetric properties of EBT films were investigated. Measurements of depth response were performed for monoenergetic fields (250 MeV/u) for different ion fluences as well as for an energy-modulated spread-out Bragg peak of 5 x 5 x 5 cm 3 in 10 cm depth. The films were positioned perpendicular to the incoming carbon ion beams. The observed quenching of the response relative to the same dose of photons was quantified by the relative efficiency. In monenenergetic beams, a relative efficiency of about 0.73 was found in the plateau, 0.4 in the peak and 0.55 in the tail region. No dependence of the relative efficiency on the ion fluence was observed well beyond the clinically used levels. This gives a constant peak to plateau ratio, which is about 1.8 times lower than that for the delivered dose. In the spread-out Bragg peak, the relative efficiency was found to decrease from 0.64 to 0.54 toward the distal end. Thus when aiming for a prediction of the film response in mixed ion beams, the efficiency of the film has to be parametrized as a function of the ion type and energy over the whole ion spectrum. In addition, the relative water-equivalent range of EBT films was measured here to be 1.291 ± 0.015.

  5. GafChromic EBT film dosimetry with flatbed CCD scanner: A novel background correction method and full dose uncertainty analysis

    International Nuclear Information System (INIS)

    Saur, Sigrun; Frengen, Jomar

    2008-01-01

    Film dosimetry using radiochromic EBT film in combination with a flatbed charge coupled device scanner is a useful method both for two-dimensional verification of intensity-modulated radiation treatment plans and for general quality assurance of treatment planning systems and linear accelerators. Unfortunately, the response over the scanner area is nonuniform, and when not corrected for, this results in a systematic error in the measured dose which is both dose and position dependent. In this study a novel method for background correction is presented. The method is based on the subtraction of a correction matrix, a matrix that is based on scans of films that are irradiated to nine dose levels in the range 0.08-2.93 Gy. Because the response of the film is dependent on the film's orientation with respect to the scanner, correction matrices for both landscape oriented and portrait oriented scans were made. In addition to the background correction method, a full dose uncertainty analysis of the film dosimetry procedure was performed. This analysis takes into account the fit uncertainty of the calibration curve, the variation in response for different film sheets, the nonuniformity after background correction, and the noise in the scanned films. The film analysis was performed for film pieces of size 16x16 cm, all with the same lot number, and all irradiations were done perpendicular onto the films. The results show that the 2-sigma dose uncertainty at 2 Gy is about 5% and 3.5% for landscape and portrait scans, respectively. The uncertainty gradually increases as the dose decreases, but at 1 Gy the 2-sigma dose uncertainty is still as good as 6% and 4% for landscape and portrait scans, respectively. The study shows that film dosimetry using GafChromic EBT film, an Epson Expression 1680 Professional scanner and a dedicated background correction technique gives precise and accurate results. For the purpose of dosimetric verification, the calculated dose distribution can

  6. GafChromic EBT film dosimetry with flatbed CCD scanner: a novel background correction method and full dose uncertainty analysis.

    Science.gov (United States)

    Saur, Sigrun; Frengen, Jomar

    2008-07-01

    Film dosimetry using radiochromic EBT film in combination with a flatbed charge coupled device scanner is a useful method both for two-dimensional verification of intensity-modulated radiation treatment plans and for general quality assurance of treatment planning systems and linear accelerators. Unfortunately, the response over the scanner area is nonuniform, and when not corrected for, this results in a systematic error in the measured dose which is both dose and position dependent. In this study a novel method for background correction is presented. The method is based on the subtraction of a correction matrix, a matrix that is based on scans of films that are irradiated to nine dose levels in the range 0.08-2.93 Gy. Because the response of the film is dependent on the film's orientation with respect to the scanner, correction matrices for both landscape oriented and portrait oriented scans were made. In addition to the background correction method, a full dose uncertainty analysis of the film dosimetry procedure was performed. This analysis takes into account the fit uncertainty of the calibration curve, the variation in response for different film sheets, the nonuniformity after background correction, and the noise in the scanned films. The film analysis was performed for film pieces of size 16 x 16 cm, all with the same lot number, and all irradiations were done perpendicular onto the films. The results show that the 2-sigma dose uncertainty at 2 Gy is about 5% and 3.5% for landscape and portrait scans, respectively. The uncertainty gradually increases as the dose decreases, but at 1 Gy the 2-sigma dose uncertainty is still as good as 6% and 4% for landscape and portrait scans, respectively. The study shows that film dosimetry using GafChromic EBT film, an Epson Expression 1680 Professional scanner and a dedicated background correction technique gives precise and accurate results. For the purpose of dosimetric verification, the calculated dose distribution

  7. Evaluation of different calibration curves QA of IMRT plans with radiochromic films; Evaluacion de diversas curvas de calibracion QA de planes de IMRT con peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Hernandez Rodriguez, J.; Martin Rincon, C.; Garcia Repiso, S.; Ramos Paheo, J. A.; Verde Velasco, J. M.; Sena Espinel, E. de

    2013-07-01

    The non-linear relationship between dose and the optical density, characteristic plates radiochromic Gafchromic EBT and EBT2, has been studied by various authors, whose publications are proposed different functional forms that fit the specific values measured curves that allow the full range of useful dose calibration. The objective of the work focuses on evaluating the influence of the use of different calibration curves in the dose measurement for quality control of IMRT treatments. (Author)

  8. Gafchromic EBT3 film dosimetry in electron beams — energy dependence and improved film read‐out

    Science.gov (United States)

    Ojala, Jarkko; Kaijaluoto, Sampsa; Jokelainen, Ilkka; Kosunen, Antti

    2016-01-01

    For megavoltage photon radiation, the fundamental dosimetry characteristics of Gafchromic EBT3 film were determined in  60Co gamma ray beam with addition of experimental and Monte Carlo (MC)‐simulated energy dependence of the film for 6 MV photon beam and 6 MeV, 9 MeV, 12 MeV, and 16 MeV electron beams in water phantom. For the film read‐out, two phase correction of scanner sensitivity was applied: a matrix correction for scanning area and dose‐dependent correction by iterative procedure. With these corrections, the uniformity of response can be improved to be within ±50 pixel values (PVs). To improve the read‐out accuracy, a procedure with flipped film orientations was established. With the method, scanner uniformity can be improved further and dust particles, scratches and/or dirt on scanner glass can be detected and eliminated. Responses from red and green channels were averaged for read‐out, which decreased the effect of noise present in values from separate channels. Since the signal level with the blue channel is considerably lower than with other channels, the signal variation due to different perturbation effects increases the noise level so that the blue channel is not recommended to be used for dose determination. However, the blue channel can be used for the detection of emulsion thickness variations for film quality evaluations with unexposed films. With electron beams ranging from 6 MeV to 16 MeV and at reference measurement conditions in water, the energy dependence of the EBT3 film is uniform within 0.5%, with uncertainties close to 1.6% (k=2). Including 6 MV photon beam and the electron beams mentioned, the energy dependence is within 1.1%. No notable differences were found between the experimental and MC‐simulated responses, indicating negligible change in intrinsic energy dependence of the EBT3 film for 6 MV photon beam and 6 MeV–16 MeV electron beams. Based on the dosimetric characteristics of the EBT3 film, the read

  9. Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry.

    Science.gov (United States)

    McCaw, Travis J; Micka, John A; DeWerd, Larry A

    2014-05-01

    Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. A film stack dosimeter was developed using Gafchromic(®) EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film

  10. Characterizing the marker-dye correction for Gafchromic(®) EBT2 film: a comparison of three analysis methods.

    Science.gov (United States)

    McCaw, Travis J; Micka, John A; Dewerd, Larry A

    2011-10-01

    Gafchromic(®) EBT2 film has a yellow marker dye incorporated into the active layer of the film that can be used to correct the film response for small variations in thickness. This work characterizes the effect of the marker-dye correction on the uniformity and uncertainty of dose measurements with EBT2 film. The effect of variations in time postexposure on the uniformity of EBT2 is also investigated. EBT2 films were used to measure the flatness of a (60)Co field to provide a high-spatial resolution evaluation of the film uniformity. As a reference, the flatness of the (60)Co field was also measured with Kodak EDR2 films. The EBT2 films were digitized with a flatbed document scanner 24, 48, and 72 h postexposure, and the images were analyzed using three methods: (1) the manufacturer-recommended marker-dye correction, (2) an in-house marker-dye correction, and (3) a net optical density (OD) measurement in the red color channel. The field flatness was calculated from orthogonal profiles through the center of the field using each analysis method, and the results were compared with the EDR2 measurements. Uncertainty was propagated through a dose calculation for each analysis method. The change in the measured field flatness for increasing times postexposure was also determined. Both marker-dye correction methods improved the field flatness measured with EBT2 film relative to the net OD method, with a maximum improvement of 1% using the manufacturer-recommended correction. However, the manufacturer-recommended correction also resulted in a dose uncertainty an order of magnitude greater than the other two methods. The in-house marker-dye correction lowered the dose uncertainty relative to the net OD method. The measured field flatness did not exhibit any unidirectional change with increasing time postexposure and showed a maximum change of 0.3%. The marker dye in EBT2 can be used to improve the response uniformity of the film. Depending on the film analysis method used

  11. Research on the formula of radiochromic film dosimeters

    International Nuclear Information System (INIS)

    Li Huazhi; Xiao Zhenhong; Lin Min; Cui Ying; Chen Kesheng; Chen Yundong; Ye Hongsheng; Lin Jingwen

    2006-10-01

    The formula of radiochromic film dosimeters was studied. Commercially available nylon was used as the matrix, while hexahydroxyethyl pararosaniline cyanide (HPR-CN) and pararosaniline cyanide (PR-CN) that are made in China and other countries were used as the dyes of the dosimeters. the performance of the thin film dosimeters made in CIAE was tested and compared with each other. The formula of the dosimeters was finally confirmed by testing its physical properties and dosimetric characteristics. (authors)

  12. Scanner Uniformity improvements for radiochromic film analysis with matt reflectance backing

    International Nuclear Information System (INIS)

    Butson, M.; Yu, P.K.N.

    2011-01-01

    Full text: A simple and reproducible method for increasing desktop scanner uniformity for the analysis of radiochromic films is presented. Scanner uniformity, especially in the non-scan direction, for transmission scanning is well known to be problematic for radiochromic film analysis and normally corrections need to be applied. These corrections are dependant on scanner coordinates and dose level applied which complicates dosimetry procedures. This study has highlighted that using reflectance scanning in combination with a matt, white backing material instead of the conventional gloss scanner finish, substantial increases in the scanner uniformity can be achieved within 90% of the scanning area. Uniformity within ±I% over the scanning area for our epsonV700 scanner tested was found. This is compared to within ±3% for reflection scanning with the gloss backing material and within ±4% for transmission scanning. The matt backing material used was simply 5 layers of standard quality white printing paper (80 g/m It was found that 5 layers was the optimal result for backing material however most of the improvements were seen with a minimum of 3 layers. Above 5 layers, no extra benefit was seen. This may eliminate the need to perform scanner corrections for position on the desktop scanners for radiochromic film dosimetry. (author)

  13. Measurements of proton energy spectra using a radiochromic film stack

    Science.gov (United States)

    Filkins, T. M.; Steidle, Jessica; Ellison, D. M.; Steidle, Jeffrey; Freeman, C. G.; Padalino, S. J.; Fiksel, G.; Regan, S. P.; Sangster, T. C.

    2014-10-01

    The energy spectrum of protons accelerated from the rear-side of a thin foil illuminated with ultra-intense laser light from the OMEGA EP laser system at the University of Rochester's Laboratory for Laser Energetics (LLE) was measured using a stack of radiochromic film (RCF). The film stack consisted of four layers of Gafchromic HD-V2 film and four layers of Gafchromic MD-V2-55 film. Aluminum foils of various thicknesses were placed between each piece of RCF in the stack. This arrangement allowed protons with energies of 30 MeV to reach the back layer of RCF in the stack. The stack was placed in the detector plane of a Thomson parabola ion energy (TPIE) spectrometer. Each piece of film in the stack was scanned using a commercially available flat-bed scanner (Epson 10000XL). The resulting optical density was converted into proton fluence using an absolute calibration of the RCF obtained at the SUNY Geneseo 1.7 MV Pelletron accelerator laboratory. In these calibration measurements, the sensitivity of the radiochromic film was measured using monoenergetic protons produced by the accelerator. Details of the analysis procedure and the resulting proton energy spectra will be presented. Funded in part by a grant from the DOE through the Laboratory for Laser Energetics.

  14. Composite depth dose measurement for total skin electron (TSE) treatments using radiochromic film

    International Nuclear Information System (INIS)

    Gamble, Lisa M; Farrell, Thomas J; Jones, Glenn W; Hayward, Joseph E

    2003-01-01

    Total skin electron (TSE) radiotherapy is routinely used to treat cutaneous T-cell lymphomas and can be implemented using a modified Stanford technique. In our centre, the composite depth dose for this technique is achieved by a combination of two patient positions per day over a three-day cycle, and two gantry angles per patient position. Due to patient morphology, underdosed regions typically occur and have historically been measured using multiple thermoluminescent dosimeters (TLDs). We show that radiochromic film can be used as a two-dimensional relative dosimeter to measure the percent depth dose in TSE radiotherapy. Composite depth dose curves were measured in a cylindrical, polystyrene phantom and compared with TLD data. Both multiple films (1 film per day) and a single film were used in order to reproduce a realistic clinical scenario. First, three individual films were used to measure the depth dose, one per treatment day, and then compared with TLD data; this comparison showed a reasonable agreement. Secondly, a single film was used to measure the dose delivered over three daily treatments and then compared with TLD data; this comparison showed good agreement throughout the depth dose, which includes doses well below 1 Gy. It will be shown that one piece of radiochromic film is sufficient to measure the composite percent depth dose for a TSE beam, hence making radiochromic film a suitable candidate for monitoring underdosed patient regions

  15. Accuracy and efficiency of published film dosimetry techniques using a flat-bed scanner and EBT3 film.

    Science.gov (United States)

    Spelleken, E; Crowe, S B; Sutherland, B; Challens, C; Kairn, T

    2018-03-01

    Gafchromic EBT3 film is widely used for patient specific quality assurance of complex treatment plans. Film dosimetry techniques commonly involve the use of transmission scanning to produce TIFF files, which are analysed using a non-linear calibration relationship between the dose and red channel net optical density (netOD). Numerous film calibration techniques featured in the literature have not been independently verified or evaluated. A range of previously published film dosimetry techniques were re-evaluated, to identify whether these methods produce better results than the commonly-used non-linear, netOD method. EBT3 film was irradiated at calibration doses between 0 and 4000 cGy and 25 pieces of film were irradiated at 200 cGy to evaluate uniformity. The film was scanned using two different scanners: The Epson Perfection V800 and the Epson Expression 10000XL. Calibration curves, uncertainty in the fit of the curve, overall uncertainty and uniformity were calculated following the methods described by the different calibration techniques. It was found that protocols based on a conventional film dosimetry technique produced results that were accurate and uniform to within 1%, while some of the unconventional techniques produced much higher uncertainties (> 25% for some techniques). Some of the uncommon methods produced reliable results when irradiated to the standard treatment doses (film analysis technique which uses transmission scanning, red colour channel analysis, netOD and a non-linear calibration curve for measuring doses up to 4000 cGy when using EBT3 film.

  16. A new radiochromic film for radiation processing

    International Nuclear Information System (INIS)

    Sidney, L.N.; Lynch, D.C.; Willett, P.S.; Englund, W.J.

    1990-01-01

    Acid-sensitive leuco dyes in combination with a chlorine-containing polymer have been used to make a new kind of radiochromic film for radiation processing. When exposed to gamma, electron beam, or high intensity ultraviolet radiation, these films undergo a color change from colorless to royal blue, fuschia, or black, depending on the dye. The dose response for gamma and electron beam radiation has been characterized using reflection and transmission spectrophotometry over an adsorbed dose range of 1 to 100 kGy. The primary features of the films include improved color stability before and after irradiation and improved moisture resistance. The response and stability of the films make them useful for indicator (qualitative) or dosimeter (quantitative) films or labels for sterilization of medical products, food (especially meat, poultry, and spices), pharmaceuticals, and cosmetics, and the crosslinking of plastics, and the curing of polymer coatings. Large pieces of the film could be used in dose mapping when setting up and validating radiation processes and medical treatments

  17. On multichannel film dosimetry with channel-independent perturbations

    International Nuclear Information System (INIS)

    Méndez, I.; Peterlin, P.; Hudej, R.; Strojnik, A.; Casar, B.

    2014-01-01

    Purpose: Different multichannel methods for film dosimetry have been proposed in the literature. Two of them are the weighted mean method and the method put forth byMicke et al. [“Multichannel film dosimetry with nonuniformity correction,” Med. Phys. 38, 2523–2534 (2011)] and Mayer et al. [“Enhanced dosimetry procedures and assessment for EBT2 radiochromic film,” Med. Phys. 39, 2147–2155 (2012)]. The purpose of this work was to compare their results and to develop a generalized channel-independent perturbations framework in which both methods enter as special cases. Methods: Four models of channel-independent perturbations were compared: weighted mean, Micke–Mayer method, uniform distribution, and truncated normal distribution. A closed-form formula to calculate film doses and the associated type B uncertainty for all four models was deduced. To evaluate the models, film dose distributions were compared with planned and measured dose distributions. At the same time, several elements of the dosimetry process were compared: film type EBT2 versus EBT3, different waiting-time windows, reflection mode versus transmission mode scanning, and planned versus measured dose distribution for film calibration and for γ-index analysis. The methods and the models described in this study are publicly accessible through IRISEU. Alpha 1.1 ( http://www.iriseu.com ). IRISEU. is a cloud computing web application for calibration and dosimetry of radiochromic films. Results: The truncated normal distribution model provided the best agreement between film and reference doses, both for calibration and γ-index verification, and proved itself superior to both the weighted mean model, which neglects correlations between the channels, and the Micke–Mayer model, whose accuracy depends on the properties of the sensitometric curves. With respect to the selection of dosimetry protocol, no significant differences were found between transmission and reflection mode scanning

  18. On multichannel film dosimetry with channel-independent perturbations

    Energy Technology Data Exchange (ETDEWEB)

    Méndez, I., E-mail: nmendez@onko-i.si; Peterlin, P.; Hudej, R.; Strojnik, A.; Casar, B. [Department of Medical Physics, Institute of Oncology Ljubljana, Zaloška cesta 2, Ljubljana 1000 (Slovenia)

    2014-01-15

    Purpose: Different multichannel methods for film dosimetry have been proposed in the literature. Two of them are the weighted mean method and the method put forth byMicke et al. [“Multichannel film dosimetry with nonuniformity correction,” Med. Phys. 38, 2523–2534 (2011)] and Mayer et al. [“Enhanced dosimetry procedures and assessment for EBT2 radiochromic film,” Med. Phys. 39, 2147–2155 (2012)]. The purpose of this work was to compare their results and to develop a generalized channel-independent perturbations framework in which both methods enter as special cases. Methods: Four models of channel-independent perturbations were compared: weighted mean, Micke–Mayer method, uniform distribution, and truncated normal distribution. A closed-form formula to calculate film doses and the associated type B uncertainty for all four models was deduced. To evaluate the models, film dose distributions were compared with planned and measured dose distributions. At the same time, several elements of the dosimetry process were compared: film type EBT2 versus EBT3, different waiting-time windows, reflection mode versus transmission mode scanning, and planned versus measured dose distribution for film calibration and for γ-index analysis. The methods and the models described in this study are publicly accessible through IRISEU. Alpha 1.1 ( http://www.iriseu.com ). IRISEU. is a cloud computing web application for calibration and dosimetry of radiochromic films. Results: The truncated normal distribution model provided the best agreement between film and reference doses, both for calibration and γ-index verification, and proved itself superior to both the weighted mean model, which neglects correlations between the channels, and the Micke–Mayer model, whose accuracy depends on the properties of the sensitometric curves. With respect to the selection of dosimetry protocol, no significant differences were found between transmission and reflection mode scanning

  19. Estimation of dose in skin through the use of radiochromic and radiographic films in patients subjected to interventional procedures

    International Nuclear Information System (INIS)

    Campos Garcia, Juan Pablo

    2014-01-01

    Radiation doses in skin of patients subjected to interventional procedures is estimated from the utilization and analysis of GAFCHROMIC® XR-RV2 radiochromic films and KODAK® X-Omat films with aid of the ImageJ software. The distribution of the radiation fields in the films is generated to obtain the distribution of dose in skin and to find peaks of dose by isodose curves using ImageJ software. The calibration curves are realized from GAFCHROMIC® XR-RV2 radiochromic films, through the use of a densitometer and two types of scanners (reflection scanner and transmission scanner). The reflection scanner has digitalized color images of 48 bit in TIFF format. The scanner transmission has digitalized in grayscale images to 16 bit in TIFF format. Each method has determined the points with maximum dose in skin. The images of the areas of regions with maximum doses are obtained of the scanner. The quantified doses are compared in the radiochromic films with the band of doses supplied by the manufacturer. The methodologies for the estimation of the doses obtained are compared of the radiochromic films with those obtained with the KODAK® X-Omat films. The procedure of obtaining of the doses is validated in patients with KODAK® X-Omat films. The doses obtained have covered a range from the 0,1Gy to 9 Gy. Radiographic films have allowed an assessment of the doses to 900 cGy due to the saturation thereof, the doses found in that range have been consistent with the doses in radiochromic films [es

  20. SU-E-T-286: Dose Verification of Spot-Scanning Proton Beam Using GafChromic EBT3 Film

    Energy Technology Data Exchange (ETDEWEB)

    Chen, C; Tang, S; Mah, D [ProCure Proton Therapy Center, Somerset, NJ (United States); Chan, M [Memorial Sloan-Kettering Cancer Center, Basking Ridge, NJ (United States)

    2015-06-15

    Purpose: Dose verification of spot-scanning proton pencil beam is performed via planar dose measurements at several depths using an ionization-chamber array, requiring repeat irradiations of each field for each depth. Here we investigate film dosimetry which has two advantages: higher resolution and efficiency from one-shot irradiation for multiple depths. Methods: Film calibration was performed using an EBT3 film at 20-cm depth of Plastic Water (CIRS, Norfolk, VA) exposed by a 10-level step wedge on a Proteus Plus proton system (IBA, Belgium). The calibration doses ranged from 25–250 cGy(RBE) for proton energies of 170–200 MeV. A uniform 1000 cm{sup 3} dose cube and a clinical prostate combined with seminal-vesicle and pelvic-nodes plan were used for this study. All treatment plans were generated in the RayStation (RaySearch Lab, Sweden). The planar doses at different depths for both cases were measured with film using triple-channel dosimetry and the MatriXX PT (IBA Dosimetry, Germany). The Gamma passing rates, dose-difference maps, and profiles of 2D planar doses measured with EBT3 film and MatriXX, versus treatment planning system (TPS) calculations were analyzed and compared using the FilmQA Pro (Ashland Inc., Bridgewater, NJ). Results: The EBT3 film measurement results matched well with the TPS calculation data with an average passing rate >95% for 2%/2mm and are comparable with the MatriXX measurements (0.7%, 1.8%, 3.8% mean differences corresponding to 3%/3mm, 3%/2mm, 2%/2mm, respectively). Overall passing rates for EBT3 films appear higher than those with MatriXX detectors. Conclusion: The energy dependence of the film response could be minimized by calibration using proton beam with mixed energies. The greater efficiency of the dose verification using GafChromic EBT3 results in a potential cost trade-off between room capacity and film cost. EBT3 film may offer distinct advantages in highly intensity-modulated fields due to its higher resolution

  1. Accelerated color development of irradiated radiochromic dye films

    International Nuclear Information System (INIS)

    Chappas, W.J.

    1981-01-01

    The radiochromic dye films developed by Chalkley and McLaughlin are quickly becoming one of the principal methods for secondary dosimetry. Their useful dose and dose rate ranges, long-term color stability, small and flexible size, and ease of reading make them ideal for spatial dose distribution measurements in the complex targets often encountered in industry. At room temperature, however, their response is slow, requiring several hours after irradiation for full color development. This work examines the effect of humidity on the film's time response and describes a method for accelerating the film's color development. By keeping the film in a controlled humidity environment or through a simple heating technique, the film can be read in minutes instead of hours after irradiation. The results are shown to be identical to those of films stored for 24 hours at room temperature

  2. Assessment of concomitant testicular dose with radiochromic film

    International Nuclear Information System (INIS)

    Fricker, Katherine; Thompson, Christine; Meyer, Juergen

    2013-01-01

    To assess the suitability of EBT2 and XRQA2 Gafchromic film for measuring low doses in the periphery of treatment fields, and to measure the accumulative concomitant dose to the contralateral testis resulting from CT imaging, pre-treatment imaging (CBCT) and seminoma radiotherapy with and without gonadal shielding. Superficial peripheral dose measurements made using EBT2 Gafchromic film on the surface of water equivalent material were compared to measurements made with an ionisation chamber in a water phantom to evaluate the suitability and accuracy of the film dosimeter for such measurements. Similarly, XRQA2 was used to measure surface doses within a kilovoltage beam and compared with ionisation chamber measurements. Gafchromic film was used to measure CT, CBCT and seminoma treatment related testicular doses on an anthropomorphic phantom. Doses were assessed for two clinical plans, both with and without gonadal shielding. Testicular doses resulting from the treatment of up to 0.83 ± 0.17 Gy were measured per treatment. Additional doses of up to 0.49 ± 0.01 and 2.35 ± 0.05 cGy were measured per CBCT and CT image, respectively. Reductions in the testicular dose in the order of 10, 36 and 78 % were observed when gonadal shielding was fitted for treatment, CT and CBCT imaging, respectively. Gafchromic film was found to be suitable for measuring dose in the periphery of treatment fields. The dose to the testis should be limited to minimise the risk of radiation related side effects. This can be achieved by using appropriate gonadal shielding, irrespective of the treatment fields employed.

  3. Radiochromic film as a radiotherapy surface-dose detector

    International Nuclear Information System (INIS)

    Butson, M.J.; Metcalfe, P.E.; Wollongong Univ., NSW; Mathur, J.N.

    1996-01-01

    Radiochromic film is shown to be a useful surface-dose detector for radiotherapy x-ray beams. Central-axis percentage surface-dose results as measured by Gafchromic film for a 6 MVp x-ray beam produced by a Varian 2100C Linac at 100 cm SSD are 16%, 25%, 35%, 41% for 10, 20, 30 and 40 cm square field sizes, respectively. Using a simple, uniform light source and a CCD camera connected to an image analysis system, quantitative 3D surface doses are accurately attainable in real time as either numerical data, a black-and-white image or a colour-enhanced image. (Author)

  4. Poster — Thur Eve — 42: Radiochromic film calibration for low-energy seed brachytherapy dose measurement

    Energy Technology Data Exchange (ETDEWEB)

    Morrison, H; Menon, G; Sloboda, R [Department of Medical Physics, Cross Cancer Institute, Edmonton, Alberta T6G 1Z2 (Canada); Department of Oncology, University of Alberta, Edmonton, Alberta T6G 2R3 (Canada)

    2014-08-15

    The purpose of this study was to investigate the accuracy of radiochromic film calibration procedures used in external beam radiotherapy when applied to I-125 brachytherapy sources delivering higher doses, and to determine any necessary modifications to achieve similar accuracy in absolute dose measurements. GafChromic EBT3 film was used to measure radiation doses upwards of 35 Gy from 6 MV, 75 kVp and (∼28 keV) I-125 photon sources. A custom phantom was used for the I-125 irradiations to obtain a larger film area with nearly constant dose to reduce the effects of film heterogeneities on the optical density (OD) measurements. RGB transmission images were obtained with an Epson 10000XL flatbed scanner, and calibration curves relating OD and dose using a rational function were determined for each colour channel and at each energy using a non-linear least square minimization method. Differences found between the 6 MV calibration curve and those for the lower energy sources are large enough that 6 MV beams should not be used to calibrate film for low-energy sources. However, differences between the 75 kVp and I-125 calibration curves were quite small; indicating that 75 kVp is a good choice. Compared with I-125 irradiation, this gives the advantages of lower type B uncertainties and markedly reduced irradiation time. To obtain high accuracy calibration for the dose range up to 35 Gy, two-segment piece-wise fitting was required. This yielded absolute dose measurement accuracy above 1 Gy of ∼2% for 75 kVp and ∼5% for I-125 seed exposures.

  5. Estimation of MSAD values in computed tomography scans using radiochromic films

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Bruno Beraldo; Teogenes Augusto da, E-mail: bbo@cdtn.br [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil); Mourao, Arnaldo Prata [Centro Federal de Educacao Tecnologica de Minas Gerais (CEFET-MG), Belo Horizonte, MG (Brazil)

    2013-03-15

    Objective: To evaluate the feasibility of using radiochromic films as an alternative dosimeter to estimate the multiple scan average dose on the basis on kerma profiles. Materials and Methods: The radiochromic films were distributed in cylinders positioned in the center and in four peripheral bores of a standard abdominal phantom utilized for computed tomography dosimetry. Results: Values for multiple scan average dose values corresponded to 13.6 {+-} 0.7, 13.5 {+-} 0.7 and 18.7 {+-} 1.0 mGy for pitch of 0.75, 1.00 and 1.50, respectively. Conclusion: In spite of results showing lower values than the reference level for radiodiagnosis (25 mGy) established by the Brazilian regulations for abdominal studies, it is suggested that there is room to optimize procedures and review the reference level for radiodiagnosis in Brazil. (author)

  6. Evolution, present state and future of the radiochromic dyeing films; Evolucion, estado actual y futuro de las peliculas de tinte radiocromico

    Energy Technology Data Exchange (ETDEWEB)

    Villarreal B, J E [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    2000-07-01

    The evolution of radiochromic films, their present state and their developing future are the object of this work. For this purpose a review of the evolution was realized and also of the present state of using the radiochromic dyeing films as tools to determine the absorbed doses distribution produces by beta emissor sources, beta-gamma, electrons and X-rays. In particular it is presented the development of radiochromic films type GafChromic that by their quality in terms of reproducibility, sensibility and high spatial resolution they have been converted in those films of greatest use so dominating market. Since one of the application fields more important of this type of films is clinical dosimetry, the more eminent applications in this area are presented, which the high resolution dosimetry that use GafChromic has been converted in a basic tool. On the other hand the scopes of this type of dosimetry and the possible development lines of dosimetry with radiochromic dyeing films are discussed. (Author)

  7. A new form of the calibration curve in radiochromic dosimetry. Properties and results

    International Nuclear Information System (INIS)

    Tamponi, Matteo; Bona, Rossana; Poggiu, Angela; Marini, Piergiorgio

    2016-01-01

    Purpose: This work describes a new form of the calibration curve for radiochromic dosimetry that depends on one fit parameter. Some results are reported to show that the new curve performs as well as those previously used and, more importantly, significantly reduces the dependence on the lot of films, the film orientation on the scanner, and the time after exposure. Methods: The form of the response curve makes use of the net optical densities ratio against the dose and has been studied by means of the Beer–Lambert law and a simple modeling of the film. The new calibration curve has been applied to EBT3 films exposed at 6 and 15 MV energy beams of linear accelerators and read-out in transmission mode by means of a flatbed color scanner. Its performance has been compared to that of two established forms of the calibration curve, which use the optical density and the net optical density against the dose. Four series of measurements with four lots of EBT3 films were used to evaluate the precision, accuracy, and dependence on the time after exposure, orientation on the scanner and lot of films. Results: The new calibration curve is roughly subject to the same dose uncertainty, about 2% (1 standard deviation), and has the same accuracy, about 1.5% (dose values between 50 and 450 cGy), as the other calibration curves when films of the same lot are used. Moreover, the new calibration curve, albeit obtained from only one lot of film, shows a good agreement with experimental data from all other lots of EBT3 films used, with an accuracy of about 2% and a relative dose precision of 2.4% (1 standard deviation). The agreement also holds for changes of the film orientation and of the time after exposure. Conclusions: The dose accuracy of this new form of the calibration curve is always equal to or better than those obtained from the two types of curves previously used. The use of the net optical densities ratio considerably reduces the dependence on the lot of films, the

  8. A new form of the calibration curve in radiochromic dosimetry. Properties and results

    Energy Technology Data Exchange (ETDEWEB)

    Tamponi, Matteo, E-mail: mtamponi@aslsassari.it; Bona, Rossana; Poggiu, Angela; Marini, Piergiorgio [Medical Physics Unit, ASL Sassari, Via Enrico de Nicola, Sassari 07100 (Italy)

    2016-07-15

    Purpose: This work describes a new form of the calibration curve for radiochromic dosimetry that depends on one fit parameter. Some results are reported to show that the new curve performs as well as those previously used and, more importantly, significantly reduces the dependence on the lot of films, the film orientation on the scanner, and the time after exposure. Methods: The form of the response curve makes use of the net optical densities ratio against the dose and has been studied by means of the Beer–Lambert law and a simple modeling of the film. The new calibration curve has been applied to EBT3 films exposed at 6 and 15 MV energy beams of linear accelerators and read-out in transmission mode by means of a flatbed color scanner. Its performance has been compared to that of two established forms of the calibration curve, which use the optical density and the net optical density against the dose. Four series of measurements with four lots of EBT3 films were used to evaluate the precision, accuracy, and dependence on the time after exposure, orientation on the scanner and lot of films. Results: The new calibration curve is roughly subject to the same dose uncertainty, about 2% (1 standard deviation), and has the same accuracy, about 1.5% (dose values between 50 and 450 cGy), as the other calibration curves when films of the same lot are used. Moreover, the new calibration curve, albeit obtained from only one lot of film, shows a good agreement with experimental data from all other lots of EBT3 films used, with an accuracy of about 2% and a relative dose precision of 2.4% (1 standard deviation). The agreement also holds for changes of the film orientation and of the time after exposure. Conclusions: The dose accuracy of this new form of the calibration curve is always equal to or better than those obtained from the two types of curves previously used. The use of the net optical densities ratio considerably reduces the dependence on the lot of films, the

  9. Calibrating page sized Gafchromic EBT3 films

    International Nuclear Information System (INIS)

    Crijns, W.; Maes, F.; Heide, U. A. van der; Van den Heuvel, F.

    2013-01-01

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittance values. The linear combination model, combined a monomer transmittance state (T 0 ) and a polymer transmittance state (T ∞ ) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the optimal balance between

  10. SU-G-BRB-14: Uncertainty of Radiochromic Film Based Relative Dose Measurements

    Energy Technology Data Exchange (ETDEWEB)

    Devic, S; Tomic, N; DeBlois, F; Seuntjens, J [McGill University, Montreal, QC (Canada); Lewis, D [RCF Consulting, LLC, Monroe, CT (United States); Aldelaijan, S [King Faisal Specialist Hospital & Research Center, Riyadh (Saudi Arabia)

    2016-06-15

    Purpose: Due to inherently non-linear dose response, measurement of relative dose distribution with radiochromic film requires measurement of absolute dose using a calibration curve following previously established reference dosimetry protocol. On the other hand, a functional form that converts the inherently non-linear dose response curve of the radiochromic film dosimetry system into linear one has been proposed recently [Devic et al, Med. Phys. 39 4850–4857 (2012)]. However, there is a question what would be the uncertainty of such measured relative dose. Methods: If the relative dose distribution is determined going through the reference dosimetry system (conversion of the response by using calibration curve into absolute dose) the total uncertainty of such determined relative dose will be calculated by summing in quadrature total uncertainties of doses measured at a given and at the reference point. On the other hand, if the relative dose is determined using linearization method, the new response variable is calculated as ζ=a(netOD)n/ln(netOD). In this case, the total uncertainty in relative dose will be calculated by summing in quadrature uncertainties for a new response function (σζ) for a given and the reference point. Results: Except at very low doses, where the measurement uncertainty dominates, the total relative dose uncertainty is less than 1% for the linear response method as compared to almost 2% uncertainty level for the reference dosimetry method. The result is not surprising having in mind that the total uncertainty of the reference dose method is dominated by the fitting uncertainty, which is mitigated in the case of linearization method. Conclusion: Linearization of the radiochromic film dose response provides a convenient and a more precise method for relative dose measurements as it does not require reference dosimetry and creation of calibration curve. However, the linearity of the newly introduced function must be verified. Dave Lewis

  11. SU-G-TeP2-06: Development of Novel Radiochromic Films for Radiotherapy Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Alqathami, M; Lee, H; Ibbott, G [UT MD Anderson Cancer Center, Houston, TX (United States); Won Choi, G [UT MD Anderson Cancer Center, Houston, TX-Texas (United States); Blencowe, A [The University of South Australia, South Australia, SA (Australia); Wen, Z [MD Anderson Cancer Center, Houston, TX (United States); Adamovics, J [Department of Chemistry and Biology, Rider University, Skillman, NJ (United States)

    2016-06-15

    Purpose: To develop and evaluate novel radiochromic films for quality assurance in radiotherapy dosimetry. Materials and Methods: Novel radiochromic film compositions were formulated using leuco crystal violet (LCV) as a reporting system and tetrabromoethane as a free radical source. The film matrix used consisted of polyurethane polymer mixed with dibutyl phthalate plasticizer (20 wt%). The concentration of the radical initiator was kept constant at 10 wt% and the concentration of the LCV dye varied (1 and 2 wt%). To ensure uniform thickness of the film, its precursors were sandwiched between two pieces of glass separated by a 1 mm gap between during the curing process. The films were cut into pieces and were irradiated with a 6 MV X-ray beam to selected doses. The change in optical density was measured using a flatbed scanner and a spectrophotometer. Results: The results showed that all film formulations exhibited a linear response with dose and an absorption maximum at ∼ 590 nm. The formulation with 2 wt% LCV was ∼ 30% more sensitive to dose than the formulation with 1 wt% LCV. Both films were very deformable. In addition, the radiochromic response of the film was found to bleach over a short period of time (few weeks) allowing the film to be reused for dose verification measurements. Conclusion: Both film formulations displayed excellent sensitivity and linearity to radiation dose and thus can be used for the 2D dosimetry of clinical megavoltage and kilovoltage X-ray beams. In addition, the thickness of the film could easily be increased allowing for their potential use as a deformable bolus material. However, thicker films would need more optimization of the manufacturing procedure to ensure consistent material uniformity and sensitivity are recommended.

  12. Determining optimum wavelength of ultraviolet rays to pre-exposure of non-uniformity error correction in Gafchromic EBT2 films

    Science.gov (United States)

    Katsuda, Toshizo; Gotanda, Rumi; Gotanda, Tatsuhiro; Akagawa, Takuya; Tanki, Nobuyoshi; Kuwano, Tadao; Noguchi, Atsushi; Yabunaka, Kouichi

    2018-03-01

    Gafchromic films have been used to measure X-ray doses in diagnostic radiology such as computed tomography. The double-exposure technique is used to correct non-uniformity error of Gafchromic EBT2 films. Because of the heel effect of diagnostic x-rays, ultraviolet A (UV-A) is intended to be used as a substitute for x-rays. When using a UV-A light-emitting diode (LED), it is necessary to determine the effective optimal UV wavelength for the active layer of Gafchromic EBT2 films. This study evaluated the relation between the increase in color density of Gafchromic EBT2 films and the UV wavelengths. First, to correct non-uniformity, a Gafchromic EBT2 film was pre-irradiated using uniform UV-A radiation for 60 min from a 72-cm distance. Second, the film was irradiated using a UV-LED with a wavelength of 353-410 nm for 60 min from a 5.3-cm distance. The maximum, minimum, and mean ± standard deviation (SD) of pixel values of the subtraction images were evaluated using 0.5 inches of a circular region of interest (ROI). The highest mean ± SD (8915.25 ± 608.86) of the pixel value was obtained at a wavelength of 375 nm. The results indicated that 375 nm is the most effective and sensitive wavelength of UV-A for Gafchromic EBT2 films and that UV-A can be used as a substitute for x-rays in the double-exposure technique.

  13. Development of radiochromic film for spatially quantitative dosimetric analysis of indirect ionizing radiation fields

    Science.gov (United States)

    Brady, Samuel Loren

    Two types of radiochromic films (RCF) were characterized for this work: EBT and XRQA film. Both films were investigated for: radiation interaction with film structure; light interaction with film structure for optimal film readout (densitometry) sensitivity; range of absorbed dose measurements; dependence of film dose measurement response as a function of changing radiation energy; fractionation and dose rate effects on film measurement response; film response sensitivity to ambient factors; and stability of measured film response with time. EBT film was shown to have the following properties: near water equivalent atomic weight (Zeff); dynamic dose range of 10 -1-102 Gy; 3% change in optical density (OD) response for a single exposure level when exposed to radiation energies from (75-18,000) kV; and best digitized using transmission densitometry. XRQA film was shown to have: a Zeff of ˜25; a 12 fold increase in sensitivity at lower photon energies for a dynamic dose range of 10-3-100 Gy, a difference of 25% in OD response when comparing 120 kV to 320 kV, and best digitized using reflective densitometry. Both XRQA and EBT films were shown to have: a temporal stability (DeltaOD) of ˜1% for t > 24 hr post film exposure for up to ˜20 days; a change in dose response of ˜0.03 mGy hr-1 when exposed to fluorescent room lighting at standard room temperature and humidity levels; a negligible dose rate and fractionation effect when operated within the optimal dose ranges; and a light wavelength dependence with dose for film readout. The flat bed scanner was chosen as the primary film digitizer due to its availability, cost, OD range, functionality (transmission and reflection scanning), and digitization speed. As a cost verses functionality comparison, the intrinsic and operational limitations were determined for two flat bed scanners. The EPSON V700 and 10000XL exhibited equal spatial and OD accuracy. The combined precision of both the scanner light sources and CCD

  14. Radiochromic dye film studies for brachytherapy applications

    International Nuclear Information System (INIS)

    Martinez-Davalos, A.; Rodriguez-Villafuerte, M.; Diaz-Perches, R.; Arzamendi-Perez, S.

    2002-01-01

    Commercial radiochromic dye films have been used in recent years to quantify absorbed dose in several medical applications. In this study we present the characterisation of the GafChromic MD-55-2 dye film, a double sensitive layer film suitable for photon irradiation in brachytherapy applications. Dose measurements were carried out with a low dose rate 137 Cs brachytherapy source, which produces very steep dose gradients in its vicinity, and therefore requires the capability of producing high spatial resolution isodose curves. Quantification of the dose rate in water per unit air kerma strength was obtained using a high-resolution transmission commercial scanner (Agfa DuoScan T1200) with the capability of digitising up to 600 x 1200 pixels per inch using 36 bits per pixel, together with optical density measurements. The Monte Carlo calculations and experimental measurements compared well in the 0-50 Gy dose interval used in this study. (author)

  15. Evaluation of a LED-based flatbed document scanner for radiochromic film dosimetry in transmission mode.

    Science.gov (United States)

    Lárraga-Gutiérrez, José Manuel; García-Garduño, Olivia Amanda; Treviño-Palacios, Carlos; Herrera-González, José Alfredo

    2018-03-01

    Flatbed scanners are the most frequently used reading instrument for radiochromic film dosimetry because its low cost, high spatial resolution, among other advantages. These scanners use a fluorescent lamp and a CCD array as light source and detector, respectively. Recently, manufacturers of flatbed scanners replaced the fluorescent lamp by light emission diodes (LED) as a light source. The goal of this work is to evaluate the performance of a commercial flatbed scanner with LED based source light for radiochromic film dosimetry. Film read out consistency, response uniformity, film-scanner sensitivity, long term stability and total dose uncertainty was evaluated. In overall, the performance of the LED flatbed scanner is comparable to that of a cold cathode fluorescent lamp (CCFL). There are important spectral differences between LED and CCFL lamps that results in a higher sensitivity of the LED scanner in the green channel. Total dose uncertainty, film response reproducibility and long-term stability of LED scanner are slightly better than those of the CCFL. However, the LED based scanner has a strong non-uniform response, up to 9%, that must be adequately corrected for radiotherapy dosimetry QA. The differences in light emission spectra between LED and CCFL lamps and its potential impact on film-scanner sensitivity suggest that the design of a dedicated flat-bed scanner with LEDs may improve sensitivity and dose uncertainty in radiochromic film dosimetry. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  16. Absolute dosimetric characterization of Gafchromic EBT3 and HDv2 films using commercial flat-bed scanners and evaluation of the scanner response function variability

    Energy Technology Data Exchange (ETDEWEB)

    Chen, S. N.; Revet, G.; Fuchs, J. [LULI-CNRS, Ecole Polytechnique, CEA: Universite Paris-Saclay, UPMC Univ Paris 06, Sorbonne Universities, F-91128 Palaiseau Cedex (France); Institute of Applied Physics, 46 Ulyanov Street, 603950 Nizhny Novgorod (Russian Federation); Gauthier, M.; Glenzer, S.; Propp, A. [SLAC National Accelerator Laboratory, 2575 Sand Hill Road, Menlo Park, California 94025 (United States); Bazalova-Carter, M. [Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8P 5C2 (Canada); Bolanos, S. [LULI-CNRS, Ecole Polytechnique, CEA: Universite Paris-Saclay, UPMC Univ Paris 06, Sorbonne Universities, F-91128 Palaiseau Cedex (France); Riquier, R. [LULI-CNRS, Ecole Polytechnique, CEA: Universite Paris-Saclay, UPMC Univ Paris 06, Sorbonne Universities, F-91128 Palaiseau Cedex (France); CEA, DAM, DIF, F-91297 Arpajon (France); Antici, P. [INRS-EMT, Varennes, J3X1S2 Québec (Canada); Morabito, A. [ELI-ALPS, ELI-HU non profit kft, Dugonics ter 13, H-6720, Szeged (Hungary); Starodubtsev, M. [Institute of Applied Physics, 46 Ulyanov Street, 603950 Nizhny Novgorod (Russian Federation)

    2016-07-15

    Radiochromic films (RCF) are commonly used in dosimetry for a wide range of radiation sources (electrons, protons, and photons) for medical, industrial, and scientific applications. They are multi-layered, which includes plastic substrate layers and sensitive layers that incorporate a radiation-sensitive dye. Quantitative dose can be retrieved by digitizing the film, provided that a prior calibration exists. Here, to calibrate the newly developed EBT3 and HDv2 RCFs from Gafchromic™, we used the Stanford Medical LINAC to deposit in the films various doses of 10 MeV photons, and by scanning the films using three independent EPSON Precision 2450 scanners, three independent EPSON V750 scanners, and two independent EPSON 11000XL scanners. The films were scanned in separate RGB channels, as well as in black and white, and film orientation was varied. We found that the green channel of the RGB scan and the grayscale channel are in fact quite consistent over the different models of the scanner, although this comes at the cost of a reduction in sensitivity (by a factor ∼2.5 compared to the red channel). To allow any user to extend the absolute calibration reported here to any other scanner, we furthermore provide a calibration curve of the EPSON 2450 scanner based on absolutely calibrated, commercially available, optical density filters.

  17. Calibrating page sized Gafchromic EBT3 films

    Energy Technology Data Exchange (ETDEWEB)

    Crijns, W.; Maes, F.; Heide, U. A. van der; Van den Heuvel, F. [Department of Radiation Oncology, University Hospitals Leuven, Herestraat 49, 3000 Leuven (Belgium); Department ESAT/PSI-Medical Image Computing, Medical Imaging Research Center, KU Leuven, Herestraat 49, 3000 Leuven (Belgium); Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Plesmanlaan 121, 1066 CX Amsterdam (Netherlands); Department of Radiation Oncology, University Hospitals Leuven, Herestraat 49, 3000 Leuven (Belgium)

    2013-01-15

    Purpose: The purpose is the development of a novel calibration method for dosimetry with Gafchromic EBT3 films. The method should be applicable for pretreatment verification of volumetric modulated arc, and intensity modulated radiotherapy. Because the exposed area on film can be large for such treatments, lateral scan errors must be taken into account. The correction for the lateral scan effect is obtained from the calibration data itself. Methods: In this work, the film measurements were modeled using their relative scan values (Transmittance, T). Inside the transmittance domain a linear combination and a parabolic lateral scan correction described the observed transmittance values. The linear combination model, combined a monomer transmittance state (T{sub 0}) and a polymer transmittance state (T{sub {infinity}}) of the film. The dose domain was associated with the observed effects in the transmittance domain through a rational calibration function. On the calibration film only simple static fields were applied and page sized films were used for calibration and measurements (treatment verification). Four different calibration setups were considered and compared with respect to dose estimation accuracy. The first (I) used a calibration table from 32 regions of interest (ROIs) spread on 4 calibration films, the second (II) used 16 ROIs spread on 2 calibration films, the third (III), and fourth (IV) used 8 ROIs spread on a single calibration film. The calibration tables of the setups I, II, and IV contained eight dose levels delivered to different positions on the films, while for setup III only four dose levels were applied. Validation was performed by irradiating film strips with known doses at two different time points over the course of a week. Accuracy of the dose response and the lateral effect correction was estimated using the dose difference and the root mean squared error (RMSE), respectively. Results: A calibration based on two films was the optimal

  18. Radiochromic film for dosimetric measurements in radiation shielding composites synthesized for applied in radiology procedures of high dose

    Energy Technology Data Exchange (ETDEWEB)

    Fontainha, C. C. P. [Universidade Federal de Minas Gerais, Departamento de Engenharia Nuclear, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil); Baptista N, A. T.; Faria, L. O., E-mail: crissia@gmail.com [Centro de Desenvolvimento da Tecnologia Nuclear / CNEN, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil)

    2015-10-15

    Full text: Medical radiology offers great benefit to patients. However, although specifics procedures of high dose, as fluoroscopy, Interventional Radiology, Computed Tomography (CT) make up a small percent of the imaging procedures, they contribute to significantly increase dose to population. The patients may suffer tissue damage. The probability of deterministic effects incidence depends on the type of procedure performed, exposure time, and the amount of applied dose at the irradiated area. Calibrated radiochromic films can identify size and distribution of the radiated fields and measure intensities of doses. Radiochromic films are sensitive for doses ranging from 0.1 to 20 c Gy and they have the same response for X-rays effective energies ranging from 20 to 100 keV. New radiation attenuators materials have been widely investigated resulting in dose reduction entrance skin dose. In this work, Bi{sub 2}O{sub 3} and ZrO{sub 2}:8 % Y{sub 2}O{sub 3} composites were obtained by mixing them with P(VDF-Tr Fe) copolymers matrix from casting method and then characterized by Ftir. Dosimetric measurements were obtained with Xr-Q A2 Gafchromic radiochromic films. In this setup, one radiochromic film is directly exposed to the X-rays beam and another one measures the attenuated beam were exposed to an absorbed dose of 10 mGy of RQR5 beam quality (70 kV X-ray beam). Under the same conditions, irradiated Xr-Q A2 films were stored and scanned measurement in order to obtain a more reliable result. The attenuation factors, evaluated by Xr-Q A2 radiochromic films, indicate that both composites are good candidates for use as patient radiation shielding in high dose medical procedures. (Author)

  19. A Study on the use of Gafchromic{sup TM} EBT3 Film for Microdosimetry by Raman Spectroscopy

    Energy Technology Data Exchange (ETDEWEB)

    Mirza, Jamal Ahmad; Heo, Taemin; Park, Hyeonsuk; Ye, Sung-Joon [Seoul National University, Seoul (Korea, Republic of)

    2015-10-15

    This concept is explained in microdosimetry, which deals with the spatial, temporal and energy spectral distributions of energy imparted in cellular and subcellular biological structures and associated biological effects. We at Seoul National University (SNU) are investigating on the use of unlaminated Gafchromic{sup TM} EBT3 film for microdosimetry. Our goal is to determine absorbed dose on EBT3 film with spatial distribution of as low as 10 micron using Raman spectroscopy. Data acquired using Raman is reproducible and temporally stable. Proper placement of film on stage and with right orientation is very important. The peak saturation problem can be avoided by decreasing laser exposure time. Raman intensity decreases with increase or decrease of depths beyond the surface active layer. Depth profile of film for each dose level maybe used to find the maximum signal, which is useful in finding the dose profile. Raman spectroscopy can be used for the determination of absorbed dose by exploiting the idea that C ≡ C and/or C=C in EBT3 film increases with increasing dose due to polymerization of diacetylene monomer.

  20. Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom

    International Nuclear Information System (INIS)

    Garcia-Garduno, Olivia A.; Larraga-Gutierrez, Jose M.; Rodriguez-Villafuerte, Mercedes; Martinez-Davalos, Arnulfo; Celis, Miguel A.

    2010-01-01

    This work reports the use of both GafChromic EBT film immersed in a water phantom and Monte Carlo (MC) simulations for small photon beam stereotactic radiosurgery dosimetry. Circularly collimated photon beams with diameters in the 4-20 mm range of a dedicated 6 MV linear accelerator (Novalis (registered) , BrainLAB, Germany) were used to perform off-axis ratios, tissue maximum ratios and total scatter factors measurements, and MC simulations. GafChromic EBT film data show an excellent agreement with MC results (<2.7%) for all measured quantities.

  1. Verification of Gamma Knife extend system based fractionated treatment planning using EBT2 film

    Energy Technology Data Exchange (ETDEWEB)

    Natanasabapathi, Gopishankar; Bisht, Raj Kishor [Gamma Knife Unit, Department of Neurosurgery, Neurosciences Centre, All India Institute of Medical Sciences, Ansari Nagar, New Delhi 110029 (India)

    2013-12-15

    Purpose: This paper presents EBT2 film verification of fractionated treatment planning with the Gamma Knife (GK) extend system, a relocatable frame system for multiple-fraction or serial multiple-session radiosurgery.Methods: A human head shaped phantom simulated the verification process for fractionated Gamma Knife treatment. Phantom preparation for Extend Frame based treatment planning involved creating a dental impression, fitting the phantom to the frame system, and acquiring a stereotactic computed tomography (CT) scan. A CT scan (Siemens, Emotion 6) of the phantom was obtained with following parameters: Tube voltage—110 kV, tube current—280 mA, pixel size—0.5 × 0.5 and 1 mm slice thickness. A treatment plan with two 8 mm collimator shots and three sectors blocking in each shot was made. Dose prescription of 4 Gy at 100% was delivered for the first fraction out of the two fractions planned. Gafchromic EBT2 film (ISP Wayne, NJ) was used as 2D verification dosimeter in this process. Films were cut and placed inside the film insert of the phantom for treatment dose delivery. Meanwhile a set of films from the same batch were exposed from 0 to 12 Gy doses for calibration purposes. An EPSON (Expression 10000 XL) scanner was used for scanning the exposed films in transparency mode. Scanned films were analyzed with inhouse written MATLAB codes.Results: Gamma index analysis of film measurement in comparison with TPS calculated dose resulted in high pass rates >90% for tolerance criteria of 1%/1 mm. The isodose overlay and linear dose profiles of film measured and computed dose distribution on sagittal and coronal plane were in close agreement.Conclusions: Through this study, the authors propose treatment verification QA method for Extend frame based fractionated Gamma Knife radiosurgery using EBT2 film.

  2. Verification of Gamma Knife extend system based fractionated treatment planning using EBT2 film

    International Nuclear Information System (INIS)

    Natanasabapathi, Gopishankar; Bisht, Raj Kishor

    2013-01-01

    Purpose: This paper presents EBT2 film verification of fractionated treatment planning with the Gamma Knife (GK) extend system, a relocatable frame system for multiple-fraction or serial multiple-session radiosurgery.Methods: A human head shaped phantom simulated the verification process for fractionated Gamma Knife treatment. Phantom preparation for Extend Frame based treatment planning involved creating a dental impression, fitting the phantom to the frame system, and acquiring a stereotactic computed tomography (CT) scan. A CT scan (Siemens, Emotion 6) of the phantom was obtained with following parameters: Tube voltage—110 kV, tube current—280 mA, pixel size—0.5 × 0.5 and 1 mm slice thickness. A treatment plan with two 8 mm collimator shots and three sectors blocking in each shot was made. Dose prescription of 4 Gy at 100% was delivered for the first fraction out of the two fractions planned. Gafchromic EBT2 film (ISP Wayne, NJ) was used as 2D verification dosimeter in this process. Films were cut and placed inside the film insert of the phantom for treatment dose delivery. Meanwhile a set of films from the same batch were exposed from 0 to 12 Gy doses for calibration purposes. An EPSON (Expression 10000 XL) scanner was used for scanning the exposed films in transparency mode. Scanned films were analyzed with inhouse written MATLAB codes.Results: Gamma index analysis of film measurement in comparison with TPS calculated dose resulted in high pass rates >90% for tolerance criteria of 1%/1 mm. The isodose overlay and linear dose profiles of film measured and computed dose distribution on sagittal and coronal plane were in close agreement.Conclusions: Through this study, the authors propose treatment verification QA method for Extend frame based fractionated Gamma Knife radiosurgery using EBT2 film

  3. SU-F-T-566: Absolute Film Dosimetry for Stereotactic Radiosurgery and Stereotactic Body Radiotherapy Quality Assurance Using Gafchromic EBT3 Films

    Energy Technology Data Exchange (ETDEWEB)

    Wen, N; Lu, S; Qin, Y; Huang, Y; Zhao, B; Liu, C; Chetty, I [Henry Ford Health System, Detroit, MI (United States)

    2016-06-15

    Purpose: To evaluate the dosimetric uncertainty associated with Gafchromic (EBT3) films and establish an absolute dosimetry protocol for Stereotactic Radiosurgery (SRS) and Stereotactic Body Radiotherapy (SBRT). Methods: EBT3 films were irradiated at each of seven different dose levels between 1 and 15 Gy with open fields, and standard deviations of dose maps were calculated at each color channel for evaluation. A scanner non-uniform response correction map was built by registering and comparing film doses to the reference diode array-based dose map delivered with the same doses. To determine the temporal dependence of EBT3 films, the average correction factors of different dose levels as a function of time were evaluated up to four days after irradiation. An integrated film dosimetry protocol was developed for dose calibration, calibration curve fitting, dose mapping, and profile/gamma analysis. Patient specific quality assurance (PSQA) was performed for 93 SRS/SBRT treatment plans. Results: The scanner response varied within 1% for the field sizes less than 5 × 5 cm{sup 2}, and up to 5% for the field sizes of 10 × 10 cm{sup 2}. The scanner correction method was able to remove visually evident, irregular detector responses found for larger field sizes. The dose response of the film changed rapidly (∼10%) in the first two hours and plateaued afterwards, ∼3% change between 2 and 24 hours. The mean uncertainties (mean of the standard deviations) were <0.5% over the dose range 1∼15Gy for all color channels for the OD response curves. The percentage of points passing the 3%/1mm gamma criteria based on absolute dose analysis, averaged over all tests, was 95.0 ± 4.2. Conclusion: We have developed an absolute film dose dosimetry protocol using EBT3 films. The overall uncertainty has been established to be approximately 1% for SRS and SBRT PSQA. The work was supported by a Research Scholar Grant, RSG-15-137-01-CCE from the American Cancer Society.

  4. Validation of the entire 2D array Octavius by radiochromic films

    International Nuclear Information System (INIS)

    Iriondo Igerabide, U.; Former Forner, A.; Otal Palacin, A.; Martin Albina, M. L.; Lozares Cordero, S.; Maneru Camara, F.; Pellejero Pellejero, S.; Miquelez Alonso, S.; Soto Prados, P. M.; Rubio Arroniz, A.

    2011-01-01

    We have validated the 2D-array Octavius set for verification of IMRT plans complete, radiating with the same angles of incidence in the patient's actual plan. This has been taken as reference measurements with radiochromic films, since they are almost isotropic response, higher resolution and we already have experience with them.

  5. Quadratic Regression-based Non-uniform Response Correction for Radiochromic Film Scanners

    International Nuclear Information System (INIS)

    Jeong, Hae Sun; Kim, Chan Hyeong; Han, Young Yih; Kum, O Yeon

    2009-01-01

    In recent years, several types of radiochromic films have been extensively used for two-dimensional dose measurements such as dosimetry in radiotherapy as well as imaging and radiation protection applications. One of the critical aspects in radiochromic film dosimetry is the accurate readout of the scanner without dose distortion. However, most of charge-coupled device (CCD) scanners used for the optical density readout of the film employ a fluorescent lamp or a coldcathode lamp as a light source, which leads to a significant amount of light scattering on the active layer of the film. Due to the effect of the light scattering, dose distortions are produced with non-uniform responses, although the dose is uniformly irradiated to the film. In order to correct the distorted doses, a method based on correction factors (CF) has been reported and used. However, the prediction of the real incident doses is difficult when the indiscreet doses are delivered to the film, since the dose correction with the CF-based method is restrictively used in case that the incident doses are already known. In a previous study, therefore, a pixel-based algorithm with linear regression was developed to correct the dose distortion of a flatbed scanner, and to estimate the initial doses. The result, however, was not very good for some cases especially when the incident dose is under approximately 100 cGy. In the present study, the problem was addressed by replacing the linear regression with the quadratic regression. The corrected doses using this method were also compared with the results of other conventional methods

  6. Measurements of the sensitivity of radiochromic film using ion beams

    Science.gov (United States)

    Steidle, J. A.; Shortino, J. P.; Ellison, D. M.; Freeman, C. G.; Sangster, T. C.

    2013-10-01

    Radiochromic film (RCF) is used in several diagnostics as a dosimeter that chromatically responds to incident particles. This response depends on the fluence, energy, and species of the incident particles. A 1.7 MV tandem Pelletron accelerator is used to create a monoenergetic ion beam which is scattered off a thin gold target onto a strip of RCF. A surface barrier detector is positioned behind a small hole in the film to measure the ion fluence on the nearby film. Once the film develops, it is scanned to examine its optical density. A response curve is acquired by fitting a three parameter formula to optical density and dose. These calibration curves can be used to help determine incident doses in a variety of situations.

  7. UNLAMINATED GAFCHROMIC EBT3 FILM FOR ULTRAVIOLET RADIATION MONITORING.

    Science.gov (United States)

    Welch, David; Randers-Pehrson, Gerhard; Spotnitz, Henry M; Brenner, David J

    2017-11-01

    Measurement of ultraviolet (UV) radiation is important for human health, especially with the expanded usage of short wavelength UV for sterilization purposes. This work examines unlaminated Gafchromic EBT3 film for UV radiation monitoring. The authors exposed the film to select wavelengths in the UV spectrum, ranging from 207 to 328 nm, and measured the change in optical density. The response of the film is wavelength dependent, and of the wavelengths tested, the film was most sensitive to 254 nm light, with measurable values as low as 10 µJ/cm2. The film shows a dose-dependent response that extends over more than four orders of magnitude. The response of the film to short wavelength UV is comparable to the daily safe exposure limits for humans, thus making it valuable as a tool for passive UV radiation monitoring. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  8. Un-laminated Gafchromic EBT3 film for ultraviolet radiation monitoring

    International Nuclear Information System (INIS)

    Welch, David; Randers-Pehrson, Gerhard; Brenner, David J.; Spotnitz, Henry M.

    2017-01-01

    Measurement of ultraviolet (UV) radiation is important for human health, especially with the expanded usage of short wavelength UV for sterilization purposes. This work examines un-laminated Gafchromic EBT3 film for UV radiation monitoring. The authors exposed the film to select wavelengths in the UV spectrum, ranging from 207 to 328 nm, and measured the change in optical density. The response of the film is wavelength dependent, and of the wavelengths tested, the film was most sensitive to 254 nm light, with measurable values as low as 10 μJ/cm 2 . The film shows a dose-dependent response that extends over more than four orders of magnitude. The response of the film to short wavelength UV is comparable to the daily safe exposure limits for humans, thus making it valuable as a tool for passive UV radiation monitoring. (authors)

  9. Technical Note: Statistical dependences between channels in radiochromic film readings. Implications in multichannel dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    González-López, Antonio, E-mail: antonio.gonzalez7@carm.es [Hospital Clínico Universitario Virgen de la Arrixaca, Ctra. Madrid-Cartagena, El Palmar, Murcia 30120 (Spain); Vera-Sánchez, Juan Antonio [Servicio de Protección Radiológica y Física Médica Hospital Universitari Sant Joan de Reus, Av. del Dr. Josep Laporte, 2, Reus, Tarragona 43204 (Spain); Ruiz-Morales, Carmen [Hospital IMED Elche, Max Planck No. 3, Elche, Alicante 03203 (Spain)

    2016-05-15

    Purpose: This note studies the statistical relationships between color channels in radiochromic film readings with flatbed scanners. The same relationships are studied for noise. Finally, their implications for multichannel film dosimetry are discussed. Methods: Radiochromic films exposed to wedged fields of 6 MV energy were read in a flatbed scanner. The joint histograms of pairs of color channels were used to obtain the joint and conditional probability density functions between channels. Then, the conditional expectations and variances of one channel given another channel were obtained. Noise was extracted from film readings by means of a multiresolution analysis. Two different dose ranges were analyzed, the first one ranging from 112 to 473 cGy and the second one from 52 to 1290 cGy. Results: For the smallest dose range, the conditional expectations of one channel given another channel can be approximated by linear functions, while the conditional variances are fairly constant. The slopes of the linear relationships between channels can be used to simplify the expression that estimates the dose by means of the multichannel method. The slopes of the linear relationships between each channel and the red one can also be interpreted as weights in the final contribution to dose estimation. However, for the largest dose range, the conditional expectations of one channel given another channel are no longer linear functions. Finally, noises in different channels were found to correlate weakly. Conclusions: Signals present in different channels of radiochromic film readings show a strong statistical dependence. By contrast, noise correlates weakly between channels. For the smallest dose range analyzed, the linear behavior between the conditional expectation of one channel given another channel can be used to simplify calculations in multichannel film dosimetry.

  10. Experimental verification of lung dose with radiochromic film: comparison with Monte Carlo simulations and commercially available treatment planning systems

    International Nuclear Information System (INIS)

    Paelinck, L; Reynaert, N; Thierens, H; Neve, W De; Wagter, C de

    2005-01-01

    The purpose of this study was to assess the absorbed dose in and around lung tissue by performing radiochromic film measurements, Monte Carlo simulations and calculations with superposition convolution algorithms. We considered a layered polystyrene phantom of 12 x 12 x 12 cm 3 containing a central cavity of 6 x 6 x 6 cm 3 filled with Gammex RMI lung-equivalent material. Two field configurations were investigated, a small 1 x 10 cm 2 field and a larger 10 x 10 cm 2 field. First, we performed Monte Carlo simulations to investigate the influence of radiochromic film itself on the measured dose distribution when the film intersects a lung-equivalent region and is oriented parallel to the central beam axis. To that end, the film and the lung-equivalent materials were modelled in detail, taking into account their specific composition. Next, measurements were performed with the film oriented both parallel and perpendicular to the central beam axis to verify the results of our Monte Carlo simulations. Finally, we digitized the phantom in two commercially available treatment planning systems, Helax-TMS version 6.1A and Pinnacle version 6.2b, and calculated the absorbed dose in the phantom with their incorporated superposition convolution algorithms to compare with the Monte Carlo simulations. Comparing Monte Carlo simulations with measurements reveals that radiochromic film is a reliable dosimeter in and around lung-equivalent regions when the film is positioned perpendicular to the central beam axis. Radiochromic film is also able to predict the absorbed dose accurately when the film is positioned parallel to the central beam axis through the lung-equivalent region. However, attention must be paid when the film is not positioned along the central beam axis, in which case the film gradually attenuates the beam and decreases the dose measured behind the cavity. This underdosage disappears by offsetting the film a few centimetres. We find deviations of about 3.6% between

  11. Experimental verification of lung dose with radiochromic film: comparison with Monte Carlo simulations and commercially available treatment planning systems

    Science.gov (United States)

    Paelinck, L.; Reynaert, N.; Thierens, H.; DeNeve, W.; DeWagter, C.

    2005-05-01

    The purpose of this study was to assess the absorbed dose in and around lung tissue by performing radiochromic film measurements, Monte Carlo simulations and calculations with superposition convolution algorithms. We considered a layered polystyrene phantom of 12 × 12 × 12 cm3 containing a central cavity of 6 × 6 × 6 cm3 filled with Gammex RMI lung-equivalent material. Two field configurations were investigated, a small 1 × 10 cm2 field and a larger 10 × 10 cm2 field. First, we performed Monte Carlo simulations to investigate the influence of radiochromic film itself on the measured dose distribution when the film intersects a lung-equivalent region and is oriented parallel to the central beam axis. To that end, the film and the lung-equivalent materials were modelled in detail, taking into account their specific composition. Next, measurements were performed with the film oriented both parallel and perpendicular to the central beam axis to verify the results of our Monte Carlo simulations. Finally, we digitized the phantom in two commercially available treatment planning systems, Helax-TMS version 6.1A and Pinnacle version 6.2b, and calculated the absorbed dose in the phantom with their incorporated superposition convolution algorithms to compare with the Monte Carlo simulations. Comparing Monte Carlo simulations with measurements reveals that radiochromic film is a reliable dosimeter in and around lung-equivalent regions when the film is positioned perpendicular to the central beam axis. Radiochromic film is also able to predict the absorbed dose accurately when the film is positioned parallel to the central beam axis through the lung-equivalent region. However, attention must be paid when the film is not positioned along the central beam axis, in which case the film gradually attenuates the beam and decreases the dose measured behind the cavity. This underdosage disappears by offsetting the film a few centimetres. We find deviations of about 3.6% between

  12. Radiochromic film measurement of spatial uniformity for a laser generated x-ray environment

    Energy Technology Data Exchange (ETDEWEB)

    Fisher, J. H.; Newlander, C. D.; Horton, R.; Fournier, K. B.; Emig, J.; Patterson, R.; Davis, J. F.; Seiler, S.; Jenkins, P. P.

    2012-10-01

    n existing x-ray source application (XRSA) test cassette was modified to hold multiple x-ray filter materials followed by two radiochromic film types (FWT-60 and HD-810 Gafchromic® film) to qualitatively characterize the spectral-spatial uniformity over the XRSA sample field of view. Multiple sets of film were examined and nominal set was determined. These initial, qualitative measurements suggest a low-energy regime (E < 3 keV) spatial anisotropy and spatial isotropy at higher energies (E > 3 keV).

  13. SU-F-T-579: Extrapolation Techniques for Small Field Dosimetry Using Gafchromic EBT3 Film

    Energy Technology Data Exchange (ETDEWEB)

    Morales, J [Chris OBrien Lifehouse, Camperdown, NSW (Australia)

    2016-06-15

    Purpose: The purpose of this project is to test an experimental approach using an extrapolation technique for Gafchromic EBT3 film for small field x-ray dosimetry. Methods: Small fields from a Novalis Tx linear accelerator with HD Multileaf Collimators with 6 MV was used. The field sizes ranged from 5 × 5 to 50 × 50 mm2 MLC fields and a range of circular cones of 4 to 30 mm2 diameters. All measurements were performed in water at an SSD of 100 cm and at a depth of 10 cm. The relative output factors (ROFs) were determined from an extrapolation technique developed to eliminate the effects of partial volume averaging in film scan by scanning films with high resolution (1200 DPI). The size of the regions of interest (ROI) was varied to produce a plot of ROFs versus ROI which was then extrapolated to zero ROI to determine the relative output factor. The results were compared with other solid state detectors with proper correction, namely, IBA SFD diode, PTW 60008 and PTW 60012 diode. Results: For the 4 mm cone, the extrapolated ROF had a value of 0.658 ± 0.014 as compared to 0.642 and 0.636 for 0.5 mm and 1 mm2 ROI analysis, respectively. This showed a change in output factor of 2.4% and 3.3% at this comparative ROI sizes. In comparison, the 25 mm cone had a difference in measured output factor of 0.3% and 0.5% between 0.5 and 1.0 mm, respectively compared to zero volume. For the fields defined by MLCs a difference of up to 2% for 5×5 mm2 was observed. Conclusion: A measureable difference can be seen in ROF based on the ROI when radiochromic film is used. Using extrapolation technique from high resolution scanning a good agreement can be achieved.

  14. Experimental study of the response of radiochromic films to proton radiation of low energy

    International Nuclear Information System (INIS)

    Mercado-Uribe, H.; Gamboa-deBuen, I.; Buenfil, A.E.; Avila, O.; Brandan, M.E.

    2009-01-01

    We have investigated the response of radiochromic films (MD-55 and HD-810) exposed to protons of 0.6 MeV. Each film is bombarded with a proton beam in an angular geometry, in such a way that the absorbed dose is related to angle. Depending on the energy and the angular fluence, the irradiated volume is total or partial. We compare the dose of these irradiated films with fully irradiated films exposed to γ radiation from a 60 Co calibrated source.

  15. Experimental comparison of profiles of acquired small fields with ionization chambers, diodes, radiochromic s and TLD films; Comparacion experimental de perfiles de campos pequenos adquiridos con camaras de ionizacion, diodos, peliculas radiocromicas y TLD

    Energy Technology Data Exchange (ETDEWEB)

    Venencia, D.; Garrigo, E. [Instituto Privado de Radioterapia, Obispo Oro 423, X5000BFI Cordoba (Argentina); Filipuzzi, M. [Instituto Balseiro, Centro Atomico Bariloche, Av. Bustillo 9500, 8400 Bariloche - Rio Negro (Argentina); Germanier, A., E-mail: devenencia@radioncologia-zunino.org [Centro de Excelencia en Productos y Procesos, Santa Maria de Punilla, 5164 Cordoba (Argentina)

    2014-08-15

    The use of radiation small fields, introduced by new techniques, can bring a considerable uncertainty in the precision of the acquired profiles, due to the conditions of lateral electronic non-equilibrium and the perturbations introduced by the detectors (volume effect and alteration of the charged particles flowing) [Das et al., 2007]. The development of new miniature detectors looks to diminish the uncertainty created by the material and the size of the sensitive volume of the dosimeter. For this reason, comparative measurements for three sizes of square field were carried out (20 mm, 10 mm and 5 mm, of side) using a detectors series: 3 ionization chambers (PTW-31003, IBA-CC04, PTW-31016), 2 diodes (PTW-60012, IBA-Sfd), thermoluminescent detectors micro-cubes of 1 mm of edge (TLD-700) and radiochromic s films EBT-3. These last two were used as reference detectors, due to their spatial high resolution and similar performance with Monte Carlo simulations [Francescon et al., 1998]. So much the thermoluminescent detectors as the radiochromic films resolved the profiles in a similar way. Both diodes responded correctly, but the rest of the detectors overestimated the gloom of the fields, which allows conclude that the used TLD (and both diodes) can resolve field sizes correctly, usually utilized in radio-surgery, without producing significant alterations in the acquired data. (author)

  16. Comparison of Epson scanner quality for radiochromic film evaluation.

    Science.gov (United States)

    Alnawaf, Hani; Yu, Peter K N; Butson, Martin

    2012-09-06

    Epson Desktop scanners have been quoted as devices which match the characteristics required for the evaluation of radiation dose exposure by radiochromic films. Specifically, models such as the 10000XL have been used successfully for image analysis and are recommended by ISP for dosimetry purposes. This note investigates and compares the scanner characteristics of three Epson desktop scanner models including the Epson 10000XL, V700, and V330. Both of the latter are substantially cheaper models capable of A4 scanning. As the price variation between the V330 and the 10000XL is 20-fold (based on Australian recommended retail price), cost savings by using the cheaper scanners may be warranted based on results. By a direct comparison of scanner uniformity and reproducibility we can evaluate the accuracy of these scanners for radiochromic film dosimetry. Results have shown that all three scanners can produce adequate scanner uniformity and reproducibility, with the inexpensive V330 producing a standard deviation variation across its landscape direction of 0.7% and 1.2% in the portrait direction (reflection mode). This is compared to the V700 in reflection mode of 0.25% and 0.5% for landscape and portrait directions, respectively, and 0.5% and 0.8% for the 10000XL. In transmission mode, the V700 is comparable in reproducibility to the 10000XL for portrait and landscape mode, whilst the V330 is only capable of scanning in the landscape direction and produces a standard deviation in this direction of 1.0% compared to 0.6% (V700) and 0.25% (10000XL). Results have shown that the V700 and 10000XL are comparable scanners in quality and accuracy with the 10000XL obviously capable of imaging over an A3 area as opposed to an A4 area for the V700. The V330 scanner produced slightly lower accuracy and quality with uncertainties approximately twice as much as the other scanners. However, the results show that the V330 is still an adequate scanner and could be used for radiation

  17. High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films.

    Science.gov (United States)

    Jaccard, Maud; Petersson, Kristoffer; Buchillier, Thierry; Germond, Jean-François; Durán, Maria Teresa; Vozenin, Marie-Catherine; Bourhis, Jean; Bochud, François O; Bailat, Claude

    2017-02-01

    The aim of this study was to assess the suitability of Gafchromic EBT3 films for reference dose measurements in the beam of a prototype high dose-per-pulse linear accelerator (linac), capable of delivering electron beams with a mean dose-rate (Ḋ m ) ranging from 0.07 to 3000 Gy/s and a dose-rate in pulse (Ḋ p ) of up to 8 × 10 6 Gy/s. To do this, we evaluated the overall uncertainties in EBT3 film dosimetry as well as the energy and dose-rate dependence of their response. Our dosimetric system was composed of EBT3 Gafchromic films in combination with a flatbed scanner and was calibrated against an ionization chamber traceable to primary standard. All sources of uncertainties in EBT3 dosimetry were carefully analyzed using irradiations at a clinical radiotherapy linac. Energy dependence was investigated with the same machine by acquiring and comparing calibration curves for three different beam energies (4, 8 and 12 MeV), for doses between 0.25 and 30 Gy. Ḋ m dependence was studied at the clinical linac by changing the pulse repetition frequency (f) of the beam in order to vary Ḋ m between 0.55 and 4.40 Gy/min, while Ḋ p dependence was probed at the prototype machine for Ḋ p ranging from 7 × 10 3 to 8 × 10 6 Gy/s. Ḋ p dependence was first determined by studying the correlation between the dose measured by films and the charge of electrons measured at the exit of the machine by an induction torus. Furthermore, we compared doses from the films to independently calibrated thermo-luminescent dosimeters (TLD) that have been reported as being dose-rate independent up to such high dose-rates. We report that uncertainty below 4% (k = 2) can be achieved in the dose range between 3 and 17 Gy. Results also demonstrated that EBT3 films did not display any detectable energy dependence for electron beam energies between 4 and 12 MeV. No Ḋ m dependence was found either. In addition, we obtained excellent consistency between films and TLDs over the entire Ḋ p

  18. The value of radiochromic film dosimetry around air cavities: experimental results and Monte Carlo simulations

    International Nuclear Information System (INIS)

    Paelinck, L; Reynaert, N; Thierens, H; Wagter, C de; Neve, W de

    2003-01-01

    In this study we investigate radiochromic film dosimetry around air cavities with particular focus on the perturbation of the dose distribution by the film when the film is parallel to the beam axis. We considered a layered polystyrene phantom containing an air cavity as a model for the air-soft tissue geometry that may occur after surgical resection of a paranasal sinus tumour. A radiochromic film type MD-55 was positioned within the phantom so that it intersected the cavity. Two phantom set-ups were examined. In the first case, the air cavity is at the centre of the phantom, thus the film is lying along the central beam axis. In the second case, the cavity and film are located 2 cm offset from the phantom centre and the central beam axis. In order to examine the influence of the film on the dose distribution and to interpret the film-measured results, Monte Carlo simulations were performed. The film was modelled rigorously to incorporate the composition and structure of the film. Two field configurations, a 1 x 10 cm 2 field and a 10 x 10 cm 2 field, were examined. The dose behind the air cavity is reduced by 6 to 7% for both field configurations when a film that intersects the cavity contains the central beam axis. This is due to the attenuation exerted by the film when photons cross the cavity. Offsetting the beam to the cavity and the film by 2 cm removes the dose reduction behind the air cavity completely. Another result was that the rebuild-up behind the cavity for the 10 x 10 cm 2 field, albeit less significant than for the 1 x 10 cm 2 field, could only be measured by the film that was placed offset with respect to the central beam axis. Although radiochromic film is approximately soft-tissue equivalent and energy independent as compared to radiographic films, care should be taken in the case of inhomogeneous phantoms when the film intersects air cavities and contains the beam central axis. Errors in dose measurement can be expected distal to the air cavity

  19. SU-E-T-30: Absorbed Doses Determined by Texture Analysis of Gafchromic EBT3 Films Using Scanning Electron Microscopy: A Feasibility Study

    International Nuclear Information System (INIS)

    Park, S; Kim, H; Ye, S

    2014-01-01

    Purpose: The texture analysis method is useful to estimate structural features of images as color, size, and shape. The study aims to determine a dose-response curve by texture analysis of Gafchromic EBT3 film images using scanning electron microscopy (SEM). Methods: The uncoated Gafchromic EBT3 films were prepared to directly scan over the active surface layer of EBT3 film using SEM. The EBT3 films were exposed at a dose range of 0 to 10 Gy using a 6 MV photon beam. The exposed film samples were SEM-scanned at 100X, 1000X, and 3000X magnifications. The four texture features (Homogeneity, Correlation, Contrast, and Energy) were calculated based on the gray level co-occurrence matrix (GLCM) derived from the SEM images at each dose. To validate a correlation between delivered doses and texture features, an R-squared value in linear regression was tested. Results: The results showed that the Correlation index was more suitable as dose indices than the other three texture features due to higher linearity and sensitivity of the dose response curves. Further the Correlation index of 3000X magnified SEM images with 9 pixel offsets had an R-squared value of 0.964. The differences between the delivered doses and the doses measured by this method were 0.9, 1.2, 0.2, and 0.2 Gy at 5, 10, 15, and 20 Gy, respectively. Conclusion: It seems to be feasible to convert micro-scale structural features of χ t χχχ he EBT3 films to absorbed doses using the texture analysis method

  20. Temperature and humidity effects on the response of radiochromic dye films

    International Nuclear Information System (INIS)

    Chappas, W.J.

    1981-01-01

    The radiochromic dye films developed by Chalkley and McLaughlin are quickly becoming one of the principal methods for secondary dosimetry. Their useful dose and dose rate ranges, long-term color stability, small and flexible size, and ease of reading make them ideal for spatial dose distribution measurements in the complex targets often encountered in research and industry. At room temperature, however, their response is slow, often requiring several hours after irradiation for full color development. This work examines the effect of humidity on the film's time response and describes two methods for accelerating the film's color development. By keeping the film in a controlled humidity environment or through a simple heating technique, the film can be read in minutes instead of hours after irradiation. The results are shown to be identical to those of films stored for 24 hours at room temperature. (author)

  1. Quality assurance in RapidArc with Alderson anthropomorphic phantom using radiochromic film in comparison to MATLAB

    International Nuclear Information System (INIS)

    Garcia, Paulo L.; Silva, Leonardo P.; Santos, Maira R.; Trindade, Cassia; Martins, Lais P.; Batista, Delano V.S.; Alves, Victor G.

    2012-01-01

    This paper presented the quality control for RapidArc using an Alderson human body phantom and radiochromic film as an alternative system to approve the treatment plan for brain tumor. Thus, it was comprised the dose distributions provided by the treatment planning system with those measured by the film radiochromic. The gamma index (Γ) analysis, to verify the acceptability of the dose distribution, was 95% of approved points, with the mostly non-compliance points in regions near the PTV’s edges. These non-compliance points may be associated to transmission blades aspects, because the regions near the edges present significant losses compared to the central areas. Also, MATLAB has proved an effective tool for that measurements and it can be used in quality assurance programs. (author)

  2. Development and clinical implementation of a global software use and analysis of radiochromic s films

    International Nuclear Information System (INIS)

    Gimeno Olmos, J.; Martinez Fernandez, J. M.; Pastor Sanchis, V.; Gonzalez Perez, V.; Guardino de la Flor, C.; Dolores Alemany, V. de los; Crispin Contreras, V.

    2011-01-01

    The aim of this study was to gather information in one solution the different processes that occur with the use of radiochromic films: calibration, reading, corrections, step-absorbed dose, dose distribution analysis and comparison with reference distributions.

  3. Characterization and review of dosimetric procedures for movies Gafchromic ETB3; Caracterizacion y revision de procedimientos dosimetricos para las peliculas Gafchromic EBT3

    Energy Technology Data Exchange (ETDEWEB)

    Perez Azorin, J. F.; Ramos Garcia, L. I.; Ortiz de Zarate Vivanco, R.; Espinosa Sola, J. M.; Celeiro Gonzalez, J.; Lupiani Castellanos, J.; Alcibar-Arechuluaga, C.; Trueba Carayo, I.

    2013-07-01

    The aim of this study is to characterize the new radiochromic EBT3 movies, presentation of dosimetry procedure followed in our department as well as comparison of our the most relevant procedure found in the literature. (Author)

  4. Computed tomography dosimeter utilizing a radiochromic film and an optical common-mode rejection: characterization and calibration of the GafChromic XRCT film

    International Nuclear Information System (INIS)

    Ohuchi, H.; Abe, M.

    2008-01-01

    Gafchromic XRCT radiochromic film is a self-developing high sensitivity radiochromic film product which can be used for assessment of delivered radiation doses which could match applications such as computed tomography (CT) dosimetry. The film automatically changes color upon irradiation changing from amber to dark greenish-black depending on the level of exposure. The absorption spectra of Gafchromic XRCT radiochromic film as measured with reflectance spectrophotometry have been investigated to analyse the dosimetry characteristics of the film. Results show two main absorption peaks produced from irradiation located at around 630 nm and 580 nm. We employed a commercially available, optical flatbed scanner for digitization of the film and image analysis software to determine the response of the XRCT films to ionizing radiation. The two dose response curves as a function of delivered dose ranging from 1.069 to 119.7 mGy for tube voltages of 80, 100, and 120 kV X-ray beams and from films scanned 24 hrs after exposure are obtained. One represents the net optical density obtained with the conventional analysis way using only red component and another shows the net reduced OD with the optical CMR scheme, which we developed, using red and green components. The measured ODs obtained with the optical CMR scheme show a good consistency among four samples and all values show an improved consistency with a second-order polynomial fit less than 1 mGy, while those with the conventional analysis exhibited a large discrepancy among four samples and did not show a consistency with a second-order polynomial fit less than 1 mGy. This result combined with its energy independence from 80 kV to 120 kV X-ray energy range provides a unique enhancement in dosimetric measurement capabilities such as the acquisition of high-spatial resolution and calibrated radiation dose profiles over currently available dosimetry films for CT applications. (author)

  5. Dose profile study in head CT scans using radiochromic films

    International Nuclear Information System (INIS)

    Ladino G, A. M.; Prata M, A.

    2016-10-01

    Diagnostic images of computed tomography generate higher doses than other methods of diagnostic radiology using X-ray beam attenuation. Clinical applications of CT have been increased by technological advances, what leads to a wide variety of scanner in the Brazilian technological pool. It has been difficult to implement dose reduction strategies because of the lack of proper guidance on computed tomography examinations. However, CT scanners allow adjusting acquisition parameter according to the patients physical profile and diagnostic application for which the scan is intended. The knowledge of the dose distribution is important because changes in image acquisition parameters may provide dose reduction. In this study, it was used a cylindrical head phantom in PMMA with 5 openings, what allows dose measurement in 5 regions. In a GE CT scanner, Discovery model of 64 channels, the central slice of the head phantom was irradiated and the absorbed doses were measured using a pencil ionization chamber. Radiochromic film strips were placed in the peripheral and in the central region of the head phantom and was performed a scan of 10 cm in the phantom central region. The scan was performed using the head scanning protocol of the radiobiology service, with a voltage of 120 kV. After scanning, the radiochromic film strips were digitalized and their digital images were used to have the dose longitudinal profiles. The dose values recorded have variation in a range of 18.66 to 23.57 mGy. In the results it was compared the dose index values obtained by the pencil chamber measurement to the dose longitudinal profiles recorded by the film strips. (Author)

  6. Dose profile study in head CT scans using radiochromic films

    Energy Technology Data Exchange (ETDEWEB)

    Ladino G, A. M.; Prata M, A., E-mail: amlgphys@gmail.com [Universidade Federal de Minas Gerais, Departamento de Engenharia Nuclear, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil)

    2016-10-15

    Diagnostic images of computed tomography generate higher doses than other methods of diagnostic radiology using X-ray beam attenuation. Clinical applications of CT have been increased by technological advances, what leads to a wide variety of scanner in the Brazilian technological pool. It has been difficult to implement dose reduction strategies because of the lack of proper guidance on computed tomography examinations. However, CT scanners allow adjusting acquisition parameter according to the patients physical profile and diagnostic application for which the scan is intended. The knowledge of the dose distribution is important because changes in image acquisition parameters may provide dose reduction. In this study, it was used a cylindrical head phantom in PMMA with 5 openings, what allows dose measurement in 5 regions. In a GE CT scanner, Discovery model of 64 channels, the central slice of the head phantom was irradiated and the absorbed doses were measured using a pencil ionization chamber. Radiochromic film strips were placed in the peripheral and in the central region of the head phantom and was performed a scan of 10 cm in the phantom central region. The scan was performed using the head scanning protocol of the radiobiology service, with a voltage of 120 kV. After scanning, the radiochromic film strips were digitalized and their digital images were used to have the dose longitudinal profiles. The dose values recorded have variation in a range of 18.66 to 23.57 mGy. In the results it was compared the dose index values obtained by the pencil chamber measurement to the dose longitudinal profiles recorded by the film strips. (Author)

  7. Small fields measurements with radiochromic films.

    Science.gov (United States)

    Gonzalez-Lopez, Antonio; Vera-Sanchez, Juan-Antonio; Lago-Martin, Jose-Domingo

    2015-01-01

    The small fields in radiotherapy are widely used due to the development of techniques such as intensity-modulated radiotherapy and stereotactic radio surgery. The measurement of the dose distributions for small fields is a challenge. A perfect dosimeter should be independent of the radiation energy and the dose rate and should have a negligible volume effect. The radiochromic (RC) film characteristics fit well to these requirements. However, the response of RC films and their digitizing processes present a significant spatial inhomogeneity problem. The present work uses a method for two-dimensional (2D) measurement with RC films based on the reduction of the spatial inhomogeneity of both the film and the film digitizing process. By means of registering and averaging several measurements of the same field, the inhomogeneities are mostly canceled. Measurements of output factors (OFs), dose profiles (in-plane and cross-plane), and 2D dose distributions are presented. The field sizes investigated are 0.5 × 0.5 cm(2), 0.7 × 0.7 cm(2), 1 × 1 cm(2), 2 × 2 cm(2), 3 × 3 cm(2), 6 × 6 cm(2), and 10 × 10 cm(2) for 6 and 15 MV photon beams. The OFs measured with the RC film are compared with the measurements carried out with a PinPoint ionization chamber (IC) and a Semiflex IC, while the measured transversal dose profiles were compared with Monte Carlo simulations. The results obtained for the OFs measurements show a good agreement with the values obtained from RC films and the PinPoint and Semiflex chambers when the field size is greater or equal than 2 × 2 cm(2). These agreements give confidence on the accuracy of the method as well as on the results obtained for smaller fields. Also, good agreement was found between the measured profiles and the Monte Carlo calculated profiles for the field size of 1 × 1 cm(2). We expect, therefore, that the presented method can be used to perform accurate measurements of small fields.

  8. Sensitivity of film measured off-axis ratios to film calibration curve using radiochromic film; Sensibilidade das razoes fora do eixo central medidas para a curva de calibracao de filmes usando filme radiocromico

    Energy Technology Data Exchange (ETDEWEB)

    Garcia-Hernandez, Diana [Universidad Nacional Autnoma de Mexico (UNAM), Mexico City (Mexico). Inst. de Fisica; Larraga-Gutierrez, Jose M. [Instituto Nacional de Neurologia y Neurocirugia, Mexico City (Mexico). Unidad de Radioneurocirugia. Lab. de Fisica Medica

    2011-07-01

    Off-axis ratios of conical beams generated with a stereotactic radiosurgery-dedicated Linac were measured with EBT2 film and stereotactic diode. The sensitivity of both full width at half maximum (FWHM) and penumbras (80-20% and 90-10%, respectively), with respect to the characteristics of the film calibration curve fit, was investigated. In all cases, penumbras resulted to be more sensitive than FWHM. However, these differences were, in general, smaller than the ones found between EBT2 reference values and the stereotactic diode measurements. The larger variation in OAR parameters was found to depend on whether the fit intersected or not the origin. A 1D gamma-index analysis showed this difference can be important in all measured conical beams. (author)

  9. A unified approach to deconvolution radiation spectra measured by radiochromic films

    CERN Document Server

    Stancic, V; Ljubenov, V

    2002-01-01

    A method for the evaluation of energy distribution of a radiation source on the basis of measured space distribution of deposited energy is proposed. The measured data were obtained by using radiochromic films. Mathematical modeling is defined as a Fredholm integral equation inversion problem. Negative solutions were treated as an additional condition expressed through undefined energy group boundaries, caused by virtue of the physical phenomenon of statistical uncertainty. Examples are given of the electron source and neutron radiation field.

  10. Evaluation of GafChromic EBT prototype B for external beam dose verification

    International Nuclear Information System (INIS)

    Todorovic, M.; Fischer, M.; Cremers, F.; Thom, E.; Schmidt, R.

    2006-01-01

    The capability of the new GafChromic EBT prototype B for external beam dose verification is investigated in this paper. First the general characteristics of this film (dose response, postirradiation coloration, influence of calibration field size) were derived using a flat-bed scanner. In the dose range from 0.1 to 8 Gy, the sensitivity of the EBT prototype B film is ten times higher than the response of the GafChromic HS, which so far was the GafChromic film with the highest sensitivity. Compared with the Kodak EDR2 film, the response of the EBT is higher by a factor of 3 in the dose range from 0.1 to 8 Gy. The GafChromic EBT almost does not show a temporal growth of the optical density and there is no influence of the chosen calibration field size on the dose response curve obtained from this data. A MatLab program was written to evaluate the two-dimensional dose distributions from treatment planning systems and GafChromic EBT film measurements. Verification of external beam therapy (SRT, IMRT) using the above-mentioned approach resulted in very small differences between the planned and the applied dose. The GafChromic EBT prototype B together with the flat-bed scanner and MatLab is a successful approach for making the advantages of the GafChromic films applicable for verification of external beam therapy

  11. Response of custom-developed radiochromic dye films after electron irradiation

    International Nuclear Information System (INIS)

    Vargas-Aburto, C.; Uribe, R.M.; McLauglin, W.L.; Dick, C.E.

    1995-01-01

    Radiochromic dye (RD) films with varying formulations have been produced in this laboratory and are being used to aid in the determination of both the absorbed dose in irradiated test materials as well as the spatial homogeneity of the electron beam used to perform the irradiations. Specifically, these films have been used during the irradiation of both photovoltaic (solar cells) and liquid crystal-based devices (light valves). However, the optical response of RD films is known to be affected by post-irradiation conditions, such as the storage time and temperature, among others. This work represents a study of the time-dependence of the response of the custom-developed RD films. The change in response has been studied for every formulation, as a function of two different post-irradiation storage temperatures (23 degrees C and 45 degrees C) for a period of six months. Results show that significant changes in the response of these films can be observed even after this extended period. These results are compared with those obtained by other authors on similar films subjected to both electron and gamma ( 60 Co) radiation

  12. On the use of Gafchromic EBT3 films for validating a commercial electron Monte Carlo dose calculation algorithm.

    Science.gov (United States)

    Chan, EuJin; Lydon, Jenny; Kron, Tomas

    2015-03-07

    This study aims to investigate the effects of oblique incidence, small field size and inhomogeneous media on the electron dose distribution, and to compare calculated (Elekta/CMS XiO) and measured results. All comparisons were done in terms of absolute dose. A new measuring method was developed for high resolution, absolute dose measurement of non-standard beams using Gafchromic® EBT3 film. A portable U-shaped holder was designed and constructed to hold EBT3 films vertically in a reproducible setup submerged in a water phantom. The experimental film method was verified with ionisation chamber measurements and agreed to within 2% or 1 mm. Agreement between XiO electron Monte Carlo (eMC) and EBT3 was within 2% or 2 mm for most standard fields and 3% or 3 mm for the non-standard fields. Larger differences were seen in the build-up region where XiO eMC overestimates dose by up to 10% for obliquely incident fields and underestimates the dose for small circular fields by up to 5% when compared to measurement. Calculations with inhomogeneous media mimicking ribs, lung and skull tissue placed at the side of the film in water agreed with measurement to within 3% or 3 mm. Gafchromic film in water proved to be a convenient high spatial resolution method to verify dose distributions from electrons in non-standard conditions including irradiation in inhomogeneous media.

  13. Dosimetry of Gamma Knife and linac-based radiosurgery using radiochromic and diode detectors

    International Nuclear Information System (INIS)

    Somigliana, A.; Borelli, S.; Zonca, G.; Pignoli, E.; Loi, G.; Marchesini, R.; Cattaneo, G.M.; Fiorino, C.; Vecchio, A. del; Calandrino, R.

    1999-01-01

    In stereotactic radiosurgery the choice of appropriate detectors, whether for absolute or relative dosimetry, is very important due to the steep dose gradient and the incomplete lateral electronic equilibrium. For both linac-based and Leksell Gamma Knife radiosurgery units, we tested the use of calibrated radiochromic film to measure absolute doses and relative dose distributions. In addition a small diode was used to estimate the relative output factors. The data obtained using radiochromic and diode detectors were compared with measurements performed with other conventional methods of dosimetry, with calculated values by treatment planning systems and with data prestored in the treatment planning system supplied by the Leksell Gamma Knife (LGK) vendor. Two stereotactic radiosurgery techniques were considered: Leksell Gamma Knife (using γ-rays from 60 Co) and linac-based radiosurgery (LR) (6 MV x-rays). Different detectors were used for both relative and absolute dosimetry: relative output factors (OFs) were estimated by using radiochromic and radiographic films and a small diode; relative dose distributions in the axial and coronal planes of a spherical polystyrene phantom were measured using radiochromic film and calculated by two different treatment planning systems (TPSs). The absolute dose at the sphere centre was measured by radiochromic film and a small ionization chamber. An accurate selection of radiochromic film was made: samples of unexposed film showing a percentage standard deviation of less than 3% were used for relative dose profiles, and for absolute dose and OF evaluations this value was reduced to 1.5%. Moreover a proper calibration curve was made for each set of measurements. With regard to absolute doses, the results obtained with the ionization chamber are in good correlation with radiochromic film-generated data, for both LGK and LR, showing a dose difference of less than 1%. The output factor evaluations, performed using different methods

  14. Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit.

    Science.gov (United States)

    Najafi, Mohsen; Geraily, Ghazale; Shirazi, Alireza; Esfahani, Mahbod; Teimouri, Javad

    2017-01-01

    In recent years, Gafchromic films are used as an advanced instrument for dosimetry systems. The EBT3 films are a new generation of Gafchromic films. Our main interest is to compare the response of the EBT3 films exposed to gamma rays provided by the Theratron 780C as a conventional radiotherapy system and the Leksell Gamma Knife as a stereotactic radiotherapy system (SRS). Both systems use Cobalt-60 sources, thus using the same energy. However, other factors such as source-to-axis distance, number of sources, dose rate, direction of irradiation, shape of phantom, the field shape of radiation, and different scatter contribution may influence the calibration curve. Calibration curves for the 2 systems were measured and plotted for doses ranging from 0 to 40 Gy at the red and green channels. The best fitting curve was obtained with the Levenberg-Marquardt algorithm. Also, the component of dose uncertainty was obtained for any calibration curve. With the best fitting curve for the EBT3 films, we can use the calibration curve to measure the absolute dose in radiation therapy. Although there is a small deviation between the 2 curves, the p-value at any channel shows no significant difference between the 2 calibration curves. Therefore, the calibration curve for each system can be the same because of minor differences. The results show that with the best fitting curve from measured data, while considering the measurement uncertainties related to them, the EBT3 calibration curve can be used to measure the unknown dose both in SRS and in conventional radiotherapy. Copyright © 2017. Published by Elsevier Inc.

  15. A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Poppinga, D., E-mail: daniela.poppinga@uni-oldenburg.de; Schoenfeld, A. A.; Poppe, B. [Medical Radiation Physics, Carl v. Ossietzky University, Oldenburg 26127, Germany and Department for Radiation Oncology, Pius Hospital, Oldenburg 26121 (Germany); Doerner, K. J. [Radiotherapy Department, General Hospital, Celle 29223 (Germany); Blanck, O. [CyberKnife Center Northern Germany, Güstrow 18273, Germany and Department for Radiation Oncology, University Clinic Schleswig-Holstein, Lübeck 23562 (Germany); Harder, D. [Medical Physics and Biophysics, Georg-August-University, Göttingen 37073 (Germany)

    2014-02-15

    Purpose: The purpose of this study is the correction of the lateral scanner artifact, i.e., the effect that, on a large homogeneously exposed EBT3 film, a flatbed scanner measures different optical densities at different positions along thex axis, the axis parallel to the elongated light source. At constant dose, the measured optical densitiy profiles along this axis have a parabolic shape with significant dose dependent curvature. Therefore, the effect is shortly called the parabola effect. The objective of the algorithm developed in this study is to correct for the parabola effect. Any optical density measured at given position x is transformed into the equivalent optical density c at the apex of the parabola and then converted into the corresponding dose via the calibration of c versus dose. Methods: For the present study EBT3 films and an Epson 10000XL scanner including transparency unit were used for the analysis of the parabola effect. The films were irradiated with 6 MV photons from an Elekta Synergy accelerator in a RW3 slab phantom. In order to quantify the effect, ten film pieces with doses graded from 0 to 20.9 Gy were sequentially scanned at eight positions along thex axis and at six positions along the z axis (the movement direction of the light source) both for the portrait and landscape film orientations. In order to test the effectiveness of the new correction algorithm, the dose profiles of an open square field and an IMRT plan were measured by EBT3 films and compared with ionization chamber and ionization chamber array measurement. Results: The parabola effect has been numerically studied over the whole measuring field of the Epson 10000XL scanner for doses up to 20.9 Gy and for both film orientations. The presented algorithm transforms any optical density at positionx into the equivalent optical density that would be measured at the same dose at the apex of the parabola. This correction method has been validated up to doses of 5.2 Gy all over the

  16. A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry

    International Nuclear Information System (INIS)

    Poppinga, D.; Schoenfeld, A. A.; Poppe, B.; Doerner, K. J.; Blanck, O.; Harder, D.

    2014-01-01

    Purpose: The purpose of this study is the correction of the lateral scanner artifact, i.e., the effect that, on a large homogeneously exposed EBT3 film, a flatbed scanner measures different optical densities at different positions along thex axis, the axis parallel to the elongated light source. At constant dose, the measured optical densitiy profiles along this axis have a parabolic shape with significant dose dependent curvature. Therefore, the effect is shortly called the parabola effect. The objective of the algorithm developed in this study is to correct for the parabola effect. Any optical density measured at given position x is transformed into the equivalent optical density c at the apex of the parabola and then converted into the corresponding dose via the calibration of c versus dose. Methods: For the present study EBT3 films and an Epson 10000XL scanner including transparency unit were used for the analysis of the parabola effect. The films were irradiated with 6 MV photons from an Elekta Synergy accelerator in a RW3 slab phantom. In order to quantify the effect, ten film pieces with doses graded from 0 to 20.9 Gy were sequentially scanned at eight positions along thex axis and at six positions along the z axis (the movement direction of the light source) both for the portrait and landscape film orientations. In order to test the effectiveness of the new correction algorithm, the dose profiles of an open square field and an IMRT plan were measured by EBT3 films and compared with ionization chamber and ionization chamber array measurement. Results: The parabola effect has been numerically studied over the whole measuring field of the Epson 10000XL scanner for doses up to 20.9 Gy and for both film orientations. The presented algorithm transforms any optical density at positionx into the equivalent optical density that would be measured at the same dose at the apex of the parabola. This correction method has been validated up to doses of 5.2 Gy all over the

  17. A new correction method serving to eliminate the parabola effect of flatbed scanners used in radiochromic film dosimetry.

    Science.gov (United States)

    Poppinga, D; Schoenfeld, A A; Doerner, K J; Blanck, O; Harder, D; Poppe, B

    2014-02-01

    The purpose of this study is the correction of the lateral scanner artifact, i.e., the effect that, on a large homogeneously exposed EBT3 film, a flatbed scanner measures different optical densities at different positions along the x axis, the axis parallel to the elongated light source. At constant dose, the measured optical density profiles along this axis have a parabolic shape with significant dose dependent curvature. Therefore, the effect is shortly called the parabola effect. The objective of the algorithm developed in this study is to correct for the parabola effect. Any optical density measured at given position x is transformed into the equivalent optical density c at the apex of the parabola and then converted into the corresponding dose via the calibration of c versus dose. For the present study EBT3 films and an Epson 10000XL scanner including transparency unit were used for the analysis of the parabola effect. The films were irradiated with 6 MV photons from an Elekta Synergy accelerator in a RW3 slab phantom. In order to quantify the effect, ten film pieces with doses graded from 0 to 20.9 Gy were sequentially scanned at eight positions along the x axis and at six positions along the z axis (the movement direction of the light source) both for the portrait and landscape film orientations. In order to test the effectiveness of the new correction algorithm, the dose profiles of an open square field and an IMRT plan were measured by EBT3 films and compared with ionization chamber and ionization chamber array measurement. The parabola effect has been numerically studied over the whole measuring field of the Epson 10000XL scanner for doses up to 20.9 Gy and for both film orientations. The presented algorithm transforms any optical density at position x into the equivalent optical density that would be measured at the same dose at the apex of the parabola. This correction method has been validated up to doses of 5.2 Gy all over the scanner bed with 2D dose

  18. Calibration of EBT2 film by the PDD method with scanner non-uniformity correction.

    Science.gov (United States)

    Chang, Liyun; Chui, Chen-Shou; Ding, Hueisch-Jy; Hwang, Ing-Ming; Ho, Sheng-Yow

    2012-09-21

    The EBT2 film together with a flatbed scanner is a convenient dosimetry QA tool for verification of clinical radiotherapy treatments. However, it suffers from a relatively high degree of uncertainty and a tedious film calibration process for every new lot of films, including cutting the films into several small pieces, exposing with different doses, restoring them back and selecting the proper region of interest (ROI) for each piece for curve fitting. In this work, we present a percentage depth dose (PDD) method that can accurately calibrate the EBT2 film together with the scanner non-uniformity correction and provide an easy way to perform film dosimetry. All films were scanned before and after the irradiation in one of the two homemade 2 mm thick acrylic frames (one portrait and the other landscape), which was located at a fixed position on the scan bed of an Epson 10 000XL scanner. After the pre-irradiated scan, the film was placed parallel to the beam central axis and sandwiched between six polystyrene plates (5 cm thick each), followed by irradiation of a 20 × 20 cm² 6 MV photon beam. Two different beams on times were used on two different films to deliver a dose to the film ranging from 32 to 320 cGy. After the post-irradiated scan, the net optical densities for a total of 235 points on the beam central axis on the films were auto-extracted and compared with the corresponding depth doses that were calculated through the measurement of a 0.6 cc farmer chamber and the related PDD table to perform the curve fitting. The portrait film location was selected for routine calibration, since the central beam axis on the film is parallel to the scanning direction, where non-uniformity correction is not needed (Ferreira et al 2009 Phys. Med. Biol. 54 1073-85). To perform the scanner non-uniformity calibration, the cross-beam profiles of the film were analysed by referencing the measured profiles from a Profiler™. Finally, to verify our method, the films were

  19. Calibration of EBT2 film by the PDD method with scanner non-uniformity correction

    International Nuclear Information System (INIS)

    Chang Liyun; Ding, Hueisch-Jy; Chui, Chen-Shou; Hwang, Ing-Ming; Ho, Sheng-Yow

    2012-01-01

    The EBT2 film together with a flatbed scanner is a convenient dosimetry QA tool for verification of clinical radiotherapy treatments. However, it suffers from a relatively high degree of uncertainty and a tedious film calibration process for every new lot of films, including cutting the films into several small pieces, exposing with different doses, restoring them back and selecting the proper region of interest (ROI) for each piece for curve fitting. In this work, we present a percentage depth dose (PDD) method that can accurately calibrate the EBT2 film together with the scanner non-uniformity correction and provide an easy way to perform film dosimetry. All films were scanned before and after the irradiation in one of the two homemade 2 mm thick acrylic frames (one portrait and the other landscape), which was located at a fixed position on the scan bed of an Epson 10 000XL scanner. After the pre-irradiated scan, the film was placed parallel to the beam central axis and sandwiched between six polystyrene plates (5 cm thick each), followed by irradiation of a 20 × 20 cm 2 6 MV photon beam. Two different beams on times were used on two different films to deliver a dose to the film ranging from 32 to 320 cGy. After the post-irradiated scan, the net optical densities for a total of 235 points on the beam central axis on the films were auto-extracted and compared with the corresponding depth doses that were calculated through the measurement of a 0.6 cc farmer chamber and the related PDD table to perform the curve fitting. The portrait film location was selected for routine calibration, since the central beam axis on the film is parallel to the scanning direction, where non-uniformity correction is not needed (Ferreira et al 2009 Phys. Med. Biol. 54 1073–85). To perform the scanner non-uniformity calibration, the cross-beam profiles of the film were analysed by referencing the measured profiles from a Profiler™. Finally, to verify our method, the films were

  20. SU-E-T-96: Energy Dependence of the New GafChromic- EBT3 Film's Dose Response-Curve.

    Science.gov (United States)

    Chiu-Tsao, S; Massillon-Jl, G; Domingo-Muñoz, I; Chan, M

    2012-06-01

    To study and compare the dose response curves of the new GafChromic EBT3 film for megavoltage and kilovoltage x-ray beams, with different spatial resolution. Two sets of EBT3 films (lot#A101711-02) were exposed to each x-ray beam (6MV, 15MV and 50kV) at 8 dose values (50-3200cGy). The megavoltage beams were calibrated per AAPM TG-51 protocol while the kilovoltage beam was calibrated following the TG-61 using an ionization chamber calibrated at NIST. Each film piece was scanned three consecutive times in the center of Epson 10000XL flatbed scanner in transmission mode, landscape orientation, 48-bit color at two separate spatial resolutions of 75 and 300 dpi. The data were analyzed using ImageJ and, for each scanned image, a region of interest (ROI) of 2×2cm 2 at the field center was selected to obtain the mean pixel value with its standard deviation in the ROI. For each energy, dose value and spatial resolution, the average netOD and its associated uncertainty were determined. The Student's t-test was performed to evaluate the statistical differences between the netOD/dose values of the three energy modalities, with different color channels and spatial resolutions. The dose response curves for the three energy modalities were compared in three color channels with 75 and 300dpi. Weak energy dependence was found. For doses above 100cGy, no statistical differences were observed between 6 and 15MV beams, regardless of spatial resolution. However, statistical differences were observed between 50kV and the megavoltage beams. The degree of energy dependence (from MV to 50kV) was found to be function of color channel, dose level and spatial resolution. The dose response curves for GafChromic EBT3 films were found to be weakly dependent on the energy of the photon beams from 6MV to 50kV. The degree of energy dependence varies with color channel, dose and spatial resolution. GafChromic EBT3 films were supplied by Ashland Corp. This work was partially supported by DGAPA

  1. Three-dimensional high dose rate dosimetry of electron beams. A combined radiochromic film, EPR and calorimetric dosimetry

    International Nuclear Information System (INIS)

    Secerov, B.; Milosavljevic, B.H.; Bacic, G.; Belgrade Univ.

    2002-01-01

    Complete text of publication follows. Aim. To examine the suitability of radiochromic film (RCF) dosimeters in determining 3D dose distribution from a pulsed electron beam source by comparing their response with alanine EPR dosimetry and calorimetry. Experimental. A FWT-60 radiochromic films (Far West Technology Inc) were used while alanine films were home made. To obtain the dose vs. penetration depth relationship, a stack of 13 films separated by aluminium plates and/or alanine films was placed perpendicular to the electron beam (Febetron, 20 ns, 1.8 MeV, 10 12 Gy/s, dose range up to 100 kGy). RC films were calibrated using 60-Co source and Fricke dosimetry. The absorbance of irradiated films was measured using 2D microdensitometry. Calorimetry was performed with a homemade quasy-adiabatic aluminum calorimeter. Results and Discussion. Microdensitometry of films (5 x 5 cm) enabled the 3D mapping of the entire radiation field with in plane resolution of 0.12 mm. The total dose for each film was obtained by image segmentation to correct for the non-linear response of films. Integrated dose for the entire stack was in good agreement (within 5%) with total absorbed energy as determined with calorimetry. The dose distribution along the beam center was determined using alanine films (1 x 1 cm) and EPR spectroscopy, and again a good agreement with the dose determined by microdensitometry of the central portion of RC films. In conclusion, the results indicate that RC films can be used for determination of 3D dose distribution even at very high dose rates

  2. Two-dimensional mapping of underdosed areas using radiochromic film for patients undergoing total skin electron beam radiotherapy

    International Nuclear Information System (INIS)

    Gamble, Lisa M.; Farrell, Thomas J.; Jones, Glenn W.; Hayward, Joseph E.

    2005-01-01

    Purpose: To demonstrate the viability of radiochromic film as an in vivo, two-dimensional dosimeter for the measurement of underdosed areas in patients undergoing total skin electron beam (TSEB) radiotherapy. The results were compared with thermoluminescent dosimeter measurements. Methods and Materials: Dosimetry results are reported for an inframammary fold of 2 patients treated using a modified version of the Stanford six-position (i.e., six-field and dual-beam) TSEB technique. The results are presented as contour plots of film optical density and percentage of dose. A linear dose profile measured from film was compared with the thermoluminescent dosimeter measurements. Results: The results showed that the percentage doses as measured by film are in good agreement with those measured by the thermoluminescent dosimeters. The isodose contour plots provided by film can be used as a two-dimensional dose map for a patient when determining the size of the supplemental patch fields. Conclusion: Radiochromic film is a viable dosimetry tool that the radiation oncologist can use to understand the surface dose heterogeneity better across complex concave regions of skin to help establish more appropriate margins to patch underdosed areas. Film could be used for patients undergoing TSEB for disorders such as mycosis fungoides or undergoing TSEB or regional skin electron beam for widespread skin metastases from breast cancer and other malignancies

  3. SU-F-I-70: Investigation of Gafchromic EBT3 Film Energy Dependence Using Proton, Photon, and Electron Beams

    International Nuclear Information System (INIS)

    Ferreira, C; Schnell, E; Ahmad, S; De La Fuente Herman, T

    2016-01-01

    Purpose: To investigate the energy dependence of Gafchromic EBT3 film over a range of clinically used proton, photon and electron energies. Methods: Proton beam energies of 117 and 204 MeV, corresponding respectively to ranges in water of 10 cm and 27 cm from a Mevion S250 double scatter system unit were used. Electron energies of 6 and 20 MeV and photon energies of 6 and 18 MV from a Varian Clinac 21EX Linac were used. Two pieces of film (5×5 cm"2) were irradiated sequentially for doses of 100, 500, and 1000 cGy for all energies and modalities. Films were placed on the central beam axis for a 10×10 cm"2 field size in the middle of spread out Bragg peak (SOBP) for proton and in respective dmax for photon and electron energies. Films were scanned on a flatbed Epson Expression 10000 XL scanner on the central region of the scanning window using 48-bit, 300 dpi, and landscape orientation after 48 hours post-irradiation of film to account for optical density (OD) stabilization. Film analysis of the red channel was performed using ImageJ 1.48v (National Institutes of Health). Results: The energy dependency of EBT3 among all energies and modalities for all doses studied was small within measurement uncertainties (1σ = ± 4.1%). The mean net OD in red channel for films receiving the same dose in the same energy modality had standard deviations within 0.9% for photons, 4.9% for electrons and 1.8% for protons. It was observed that film pieces were activated during proton irradiation, e.g., 7 mR/hr at surface after 30 minutes of irradiation, lasting for 2 hours post irradiation. Conclusion: EBT3 energy dependency was evaluated for clinically used proton, photon, and electron energies. The film self-activation may have contributed to fog and negligible dose.

  4. SU-F-I-70: Investigation of Gafchromic EBT3 Film Energy Dependence Using Proton, Photon, and Electron Beams

    Energy Technology Data Exchange (ETDEWEB)

    Ferreira, C; Schnell, E; Ahmad, S; De La Fuente Herman, T [University of Oklahoma Health Science Center, Oklahoma City, OK (United States)

    2016-06-15

    Purpose: To investigate the energy dependence of Gafchromic EBT3 film over a range of clinically used proton, photon and electron energies. Methods: Proton beam energies of 117 and 204 MeV, corresponding respectively to ranges in water of 10 cm and 27 cm from a Mevion S250 double scatter system unit were used. Electron energies of 6 and 20 MeV and photon energies of 6 and 18 MV from a Varian Clinac 21EX Linac were used. Two pieces of film (5×5 cm{sup 2}) were irradiated sequentially for doses of 100, 500, and 1000 cGy for all energies and modalities. Films were placed on the central beam axis for a 10×10 cm{sup 2} field size in the middle of spread out Bragg peak (SOBP) for proton and in respective dmax for photon and electron energies. Films were scanned on a flatbed Epson Expression 10000 XL scanner on the central region of the scanning window using 48-bit, 300 dpi, and landscape orientation after 48 hours post-irradiation of film to account for optical density (OD) stabilization. Film analysis of the red channel was performed using ImageJ 1.48v (National Institutes of Health). Results: The energy dependency of EBT3 among all energies and modalities for all doses studied was small within measurement uncertainties (1σ = ± 4.1%). The mean net OD in red channel for films receiving the same dose in the same energy modality had standard deviations within 0.9% for photons, 4.9% for electrons and 1.8% for protons. It was observed that film pieces were activated during proton irradiation, e.g., 7 mR/hr at surface after 30 minutes of irradiation, lasting for 2 hours post irradiation. Conclusion: EBT3 energy dependency was evaluated for clinically used proton, photon, and electron energies. The film self-activation may have contributed to fog and negligible dose.

  5. SU-E-T-497: Semi-Automated in Vivo Radiochromic Film Dosimetry Using a Novel Image Processing Algorithm

    International Nuclear Information System (INIS)

    Reyhan, M; Yue, N

    2014-01-01

    Purpose: To validate an automated image processing algorithm designed to detect the center of radiochromic film used for in vivo film dosimetry against the current gold standard of manual selection. Methods: An image processing algorithm was developed to automatically select the region of interest (ROI) in *.tiff images that contain multiple pieces of radiochromic film (0.5x1.3cm 2 ). After a user has linked a calibration file to the processing algorithm and selected a *.tiff file for processing, an ROI is automatically detected for all films by a combination of thresholding and erosion, which removes edges and any additional markings for orientation. Calibration is applied to the mean pixel values from the ROIs and a *.tiff image is output displaying the original image with an overlay of the ROIs and the measured doses. Validation of the algorithm was determined by comparing in vivo dose determined using the current gold standard (manually drawn ROIs) versus automated ROIs for n=420 scanned films. Bland-Altman analysis, paired t-test, and linear regression were performed to demonstrate agreement between the processes. Results: The measured doses ranged from 0.2-886.6cGy. Bland-Altman analysis of the two techniques (automatic minus manual) revealed a bias of -0.28cGy and a 95% confidence interval of (5.5cGy,-6.1cGy). These values demonstrate excellent agreement between the two techniques. Paired t-test results showed no statistical differences between the two techniques, p=0.98. Linear regression with a forced zero intercept demonstrated that Automatic=0.997*Manual, with a Pearson correlation coefficient of 0.999. The minimal differences between the two techniques may be explained by the fact that the hand drawn ROIs were not identical to the automatically selected ones. The average processing time was 6.7seconds in Matlab on an IntelCore2Duo processor. Conclusion: An automated image processing algorithm has been developed and validated, which will help minimize

  6. Extraction of electron beam dose parameters from EBT2 film data scored in a mini phantom.

    Science.gov (United States)

    O'Reilly, Dedri; Smit, Cobus J L; du Plessis, Freek C P

    2013-09-01

    Quality assurance of medical linear accelerators includes dosimetric parameter measurement of therapeutic electron beams e.g. relative dose at a depth of 80% (R₈₀). This parameter must be within a tolerance of 0.2 cm of the declared value. Cumbersome water tank measurements can be regarded as a benchmark to measure electron depth dose curves. A mini-phantom was designed and built, in which a strip of GAFCHROMIC® EBT2 film could be encased tightly for electron beam depth dose measurement. Depth dose data were measured for an ELEKTA Sl25 MLC, ELEKTA Precise, and ELEKTA Synergy (Elekta Oncology Systems, Crawley, UK) machines. The electron beam energy range was between 4 and 22 MeV among the machines. A 10 × 10 cm² electron applicator with 95 cm source-surface-distance was used on all the machines. 24 h after irradiation, the EBT2 film strips were scanned on Canon CanoScan N670U scanner. Afterwards, the data were analysed with in-house developed software that entailed optical density to dose conversion, and optimal fitting of the PDD data to de-noise the raw data. From the PDD data R₈₀ values were solved for and compared with acceptance values. A series of tests were also carried out to validate the use of the scanner for film Dosimetry. These tests are presented in this study. It was found that this method of R₈₀ evaluation was reliable with good agreement with benchmark water tank measurements using a commercial parallel plate ionization chamber as the radiation detector. The EBT2 film data yielded R₈₀ values that were on average 0.06 cm different from benchmark water tank measured R₈₀ values.

  7. SU-E-T-121: Dosimetric Characterization of Gafchromic Film EBT3 Using Vidar DosimetryPro Advantage RED and EPSON Expression 10000XL Scanners

    International Nuclear Information System (INIS)

    Medina, L; Adrada, A; Filipuzzi, M; Garrigo, E; Venencia, C

    2014-01-01

    Purpose: The purpose of this paper is to characterize EBT3 using two types of scanner, analyzing the factors of influence of each dosimetry system. Methods: The film used in this study was GAFCHROMIC EBT3, the films were exposed at a dose range between 0Gy a 9Gy in a solid water phantom, SSD=100cm, 5cm depth and perpendicularly to the 6MV photon beam generated by a Novalis TX linear accelerator equipped with an HDMLC. A Farmer type ion chamber TN30013 (PTW) was used to determine the dose delivered to the film. The films were digitized with a scanner EPSON expression 10000XL and the VIDAR DosimetryPro Adventage RED. Software RIT113v6.1 was used for construction of the calibration curve and analysis. The film characteristics investigated were: response at different dose levels, sensitivity to orientation and side and resolution through the results of the spatial response function by analyzing a step pattern. Additionally, 20 IMRT treatment fields were measured with both scanner and compared with calculated dose using gamma index analysis (3%-3mm). Results: The OD obtained for dose level 2Gy in the orientation portrait of the film on the scanner EPSON is (0,222±0,19) and for Vidar RED (0,252±0,10) and landscape is for EPSON (0,211±0,25) and for Vidar RED (0,250±0,11) . The orientation dependence with respect to film side is about 0,09% for EPSON and about 0.03% for VIDAR. The spatial response function increase in response to the Gaussian function FWHM EPSON scanner (0.18mm) compared with VIDAR scanner function (less than 0.06mm) was observed. We analyzed 20 total plan dose distributions the number of pixels with gamma>1 (3%-3mm) was 0.7%±1.2 [0.1%; 2.82%] for EBT3-VIDAR y 2%±2.9 [0.2%; 3.5%] for EBT3-EPSON. Conclusion: VIDAR scanner shows better sensitivity. EBT3 film shows a different response between portrait and landscape orientation. Step pattern is better reproduce by VIDAR scanner

  8. SU-E-T-121: Dosimetric Characterization of Gafchromic Film EBT3 Using Vidar DosimetryPro Advantage RED and EPSON Expression 10000XL Scanners

    Energy Technology Data Exchange (ETDEWEB)

    Medina, L; Adrada, A; Filipuzzi, M; Garrigo, E; Venencia, C [Instituto Privado de Radioterapia - Fundacion Marie Curie, Cordoba (Argentina)

    2014-06-01

    Purpose: The purpose of this paper is to characterize EBT3 using two types of scanner, analyzing the factors of influence of each dosimetry system. Methods: The film used in this study was GAFCHROMIC EBT3, the films were exposed at a dose range between 0Gy a 9Gy in a solid water phantom, SSD=100cm, 5cm depth and perpendicularly to the 6MV photon beam generated by a Novalis TX linear accelerator equipped with an HDMLC. A Farmer type ion chamber TN30013 (PTW) was used to determine the dose delivered to the film. The films were digitized with a scanner EPSON expression 10000XL and the VIDAR DosimetryPro Adventage RED. Software RIT113v6.1 was used for construction of the calibration curve and analysis. The film characteristics investigated were: response at different dose levels, sensitivity to orientation and side and resolution through the results of the spatial response function by analyzing a step pattern. Additionally, 20 IMRT treatment fields were measured with both scanner and compared with calculated dose using gamma index analysis (3%-3mm). Results: The OD obtained for dose level 2Gy in the orientation portrait of the film on the scanner EPSON is (0,222±0,19) and for Vidar RED (0,252±0,10) and landscape is for EPSON (0,211±0,25) and for Vidar RED (0,250±0,11) . The orientation dependence with respect to film side is about 0,09% for EPSON and about 0.03% for VIDAR. The spatial response function increase in response to the Gaussian function FWHM EPSON scanner (0.18mm) compared with VIDAR scanner function (less than 0.06mm) was observed. We analyzed 20 total plan dose distributions the number of pixels with gamma>1 (3%-3mm) was 0.7%±1.2 [0.1%; 2.82%] for EBT3-VIDAR y 2%±2.9 [0.2%; 3.5%] for EBT3-EPSON. Conclusion: VIDAR scanner shows better sensitivity. EBT3 film shows a different response between portrait and landscape orientation. Step pattern is better reproduce by VIDAR scanner.

  9. Engineering design of a toroidal divertor for the EBT-S fusion device. Final report, Phase II. EBT-S divertor project

    International Nuclear Information System (INIS)

    Mai, L.P.; Malick, F.S.

    1981-01-01

    The mechanical, structural, thermal, electrical, and vacuum design of a magnetic toroidal divertor system for the Elmo Bumpy Torus (EBT-S) is presented. The EBT-S is a toroidal magnetic fusion device located at the ORNL that operates under steady state conditions. The engineering of the divertor was performed during the second of three phases of a program aimed at the selection, design, fabrication, and installation of a magnetic divertor for EBT-S. The magnetic analysis of the toroidal divertor was performed during Phase I of the program and has been reported in a separate document. In addition to the details of the divertor design, the modest modifications that are required to the EBT-S device and facility to accommodate the divertor system are presented

  10. Effects of read-out light sources and ambient light on radiochromic film

    International Nuclear Information System (INIS)

    Butson, Martin J.; Yu, Peter K.N.; Metcalfe, Peter E.

    1998-01-01

    Both read-out light sources and ambient light sources can produce a marked effect on coloration of radiochromic film. Fluorescent, helium neon laser, light emitting diode (LED) and incandescent read-out light sources produce an equivalent dose coloration of 660 cGy h -1 , 4.3 cGy h -1 , 1.7 cGy h -1 and 2.6 cGy h -1 respectively. Direct sunlight, fluorescent light and incandescent ambient light produce an equivalent dose coloration of 30 cGy h -1 , 18 cGy h -1 and 0 cGy h -1 respectively. Continuously on, fluorescent light sources should not be used for film optical density evaluation and minimal exposure to any light source will increase the accuracy of results. (author)

  11. Radiochromic film sensitivity calibrations using ion beams from a Pelletron accelerator

    Science.gov (United States)

    Filkins, T. M.; Steidle, Jessica; Ward, R. J.; Freeman, C. G.; Padalino, S. J.; Regan, S. P.; Sangster, T. C.

    2015-11-01

    Radiochromic film (RCF) is a transparent detector film that permanently changes color following exposure to ionizing radiation. The optical density of the film increases with increasing absorbed dose. RCF is convenient to use because it requires no chemical processing and can be scanned using commercially available document scanners. RCF is used frequently in medical applications, but is also used in a variety of diagnostics in high energy density physics. The film consists of a single or double layer of radiation-sensitive organic microcrystal monomers placed onto a polyester backing. GafchromicTM manufactures a large number of different types of RCF, and new types of film frequently replace older products. In this study, the sensitivity of several types of RCF to ion beams of different energies was measured. Ion beams produced by the SUNY Geneseo 1.7 MV Pelletron accelerator were directed into a target chamber where they scattered off of a gold foil. A sample of RCF was exposed to the scattered ions. The fluence of incident particles on the film was measured using a surface barrier detector. Results of these calibrations will be presented. This work was funded in part by a grant from the DOE through the Laboratory for Laser Energetics.

  12. SU-E-T-231: Measurements of Gold Nanoparticle-Mediated Proton Dose Enhancement Due to Particle-Induced X-Ray Emission and Activation Products Using Radiochromic Films and CdTe Detector

    International Nuclear Information System (INIS)

    Cho, J; Cho, S; Manohar, N; Krishnan, S

    2014-01-01

    Purpose: There have been several reports of enhanced cell-killing and tumor regression when tumor cells and mouse tumors were loaded with gold nanoparticles (GNPs) prior to proton irradiation. While particle-induced xray emission (PIXE), Auger electrons, secondary electrons, free radicals, and biological effects have been suggested as potential mechanisms responsible for the observed GNP-mediated dose enhancement/radiosensitization, there is a lack of quantitative analysis regarding the contribution from each mechanism. Here, we report our experimental effort to quantify some of these effects. Methods: 5-cm-long cylindrical plastic vials were filled with 1.8 mL of either water or water mixed with cylindrical GNPs at the same gold concentration (0.3 mg Au/g) as used in previous animal studies. A piece of EBT2 radiochromic film (30-µm active-layer sandwiched between 80/175-µm outer-layers) was inserted along the long axis of each vial and used to measure dose enhancement due to PIXE from GNPs. Vials were placed at center-of-modulation (COM) and 3-cm up-/down-stream from COM and irradiated with 5 different doses (2–10 Gy) using 10-cm-SOBP 160-MeV protons. After irradiation, films were cleaned and read to determine the delivered dose. A vial containing spherical GNPs (20 mg Au/g) was also irradiated, and gamma-rays from activation products were measured using a cadmium-telluride (CdTe) detector. Results: Film measurements showed no significant dose enhancement beyond the experimental uncertainty (∼2%). There was a detectable activation product from GNPs, but it appeared to contribute to dose enhancement minimally (<0.01%). Conclusion: Considering the composition of EBT2 film, it can be inferred that gold characteristic x-rays from PIXE and their secondary electrons make insignificant contribution to dose enhancement. The current investigation also suggests negligible dose enhancement due to activation products. Thus, previously-reported GNP-mediated proton dose

  13. Radiochromic film dosimetry of contrast-enhanced radiotherapy (CERT)

    International Nuclear Information System (INIS)

    Morris, Kevin N; Weil, Michael D; Malzbender, Rainer

    2006-01-01

    Contrast-enhanced radiotherapy (CERT) employs elevated concentrations of high-Z contrast media in targets to dramatically increase the absorbed dose of radiation relative to the surrounding tissues. However, it is difficult to measure the dose enhancement with routine clinical instruments because the photoelectrons and Auger electrons produced by the interaction of kilovoltage x-rays with the contrast agent travel extremely short distances. We have developed a technique utilizing unlaminated radiochromic film to measure the maximum dose enhancement factor attainable in solutions of contrast agent with iodine concentrations ranging from 0% to 37% (w/v). The films were also used to simultaneously measure the dose enhancement and beam attenuation through solutions containing 0%, 5% and 10% iodine (w/v). These depth-dose measurements were made in phantoms representing 3 cm targets located 0 and 4.8 cm deep in acrylic. Higher concentrations of contrast yielded greater dose enhancement and target-to-surface dose ratios, but caused more rapid attenuation of the beam as it traversed the target. The dose distribution across the entire target could be improved by employing multiple beams. These results, using concentrations of contrast that are clinically achievable by intratumoural injection, illustrate the potential of CERT for the specific delivery of high doses of radiation to targets at depth

  14. Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique

    International Nuclear Information System (INIS)

    Chang, Liyun; Ho, Sheng-Yow; Lee, Tsair-Fwu; Yeh, Shyh-An; Ding, Hueisch-Jy; Chen, Pang-Yu

    2015-01-01

    Purpose: Ashland Inc. EBT2 and EBT3 films are widely used in quality assurance for radiation therapy; however, there remains a relatively high degree of uncertainty [B. Hartmann, M. Martisikova, and O. Jakel, “Homogeneity of Gafchromic EBT2 film,” Med. Phys. 37, 1753–1756 (2010)]. Micke et al. (2011) recently improved the spatial homogeneity using all color channels of a flatbed scanner; however, van Hoof et al. (2012) pointed out that the corrected nonuniformity still requires further investigation for larger fields. To reduce the calibration errors and the uncertainty, the authors propose a new red-channel percentage-depth-dose method in combination with a modified three-channel technique. Methods: For the ease of comparison, the EBT2 film image used in the authors’ previous study (2012) was reanalyzed using different approaches. Photon beams of 6-MV were delivered to two different films at two different beam on times, resulting in the absorption doses of ranging from approximately 30 to 300 cGy at the vertical midline of the film, which was set to be coincident with the central axis of the beam. The film was tightly sandwiched in a 30 3 -cm 3 polystyrene phantom, and the pixel values for red, green, and blue channels were extracted from 234 points on the central axis of the beam and compared with the corresponding depth doses. The film was first calibrated using the multichannel method proposed by Micke et al. (2010), accounting for nonuniformities in the scanner. After eliminating the scanner and dose-independent nonuniformities, the film was recalibrated via the dose-dependent optical density of the red channel and fitted to a power function. This calibration was verified via comparisons of the dose profiles extracted from the films, where three were exposed to a 60° physical wedge field and three were exposed to composite fields, and all of which were measured in a water phantom. A correction for optical attenuation was implemented, and treatment

  15. Absolute dose measurement Gafchromic R EBT2 movies. Case Study of Kaposis sarcoma; Medida de dosis absoluta con peliculas Gafchromic EBT2. Caso practico de un sarcoma de Kaposi

    Energy Technology Data Exchange (ETDEWEB)

    Pereira, L.; Moral, F. del; Meilan, E.; Azevedo Gomes, J. C. de; Tejeiro Garcia, A. G.; Andrade Alvarez, B.; Vazquez, J.; Nieto, I.; Medal, D.; Lopez Medina, A.; Francisco, S.; Salgado, M.; Munoz, V.

    2011-07-01

    Because of its high spatial resolution, low energy dependence and good response over a wide energy range, EBT2 Gafchromic films are widely used in many applications in radiotherapy for measuring relative dose. Despite being the most common use can be used to measure absolute dose. This text is an example of using films as EBT2 for in vivo absolute dose in a Kaposis sarcoma.

  16. Response of radiochromic dye films to low energy heavy charged particles

    CERN Document Server

    Buenfil, A E; Gamboa-Debuen, I; Aviles, P; Avila, O; Olvera, C; Robledo, R; Rodriguez-Ponce, M; Mercado-Uribe, H; Rodriguez-Villafuerte, M; Brandan, M E

    2002-01-01

    We have studied the possible use of radiochromic dye films (RCF) as heavy charged particle dosemeters. We present the results of irradiating two commercial RCF (GafChromic HD-810 and MD-55-1) with 1.5, 2.9 and 4.4 MeV protons, 1.4, 2.8, 4.7, 5.9, 6.8 MeV sup 4 He ions and 8.5 and 12.4 MeV sup 1 sup 2 C ions, at proton doses from about 1 Gy up to 3 kGy, helium ions doses from 3 Gy to 5 kGy and carbon ion doses from 30 Gy to 20 kGy. The films were scanned and digitized using commercial equipment. For a given particle, the response per unit dose at different energies indicates an energy dependence of the sensitivity, which is discussed. Comparison was made for the use of a standard spectrophotometer to obtain optical density readings versus a white light scanner.

  17. Dosimetry using Gafchromic XR-RV2 radiochromic films in interventional radiology

    International Nuclear Information System (INIS)

    Neocleous, A.; Yakoumakis, E.; Gialousis, G.; Dimitriadis, A.; Yakoumakis, N.; Georgiou, E.

    2011-01-01

    Patient dose measurements of local entrance dose to the skin have been carried out using radiochromic film (Gafchromic XR-RV2) in a sample of interventional procedures. The major aim of the work was to measure patient entrance dose from such examinations using Gafchromic XR-RV2. Forty-five various interventional procedures (including nephrostomies and urinary stenting, biliary stenting and percutaneous transhepatic biliary drainage (PTBD) and aorta stent grafting) were evaluated. Maximum entrance doses were 537±119 mGy in nephrostomies, 943±631 mGy in biliary stenting and PTBD and 2425±569 mGy in aorta stent grafting. Results indicate that all patients undergoing aorta stent grafting received skin dose above 1500 mGy, which means that there is an increasing potential to suffer radiation-induced skin injuries. The film provides dose mapping, the position of the skin area with highest dose and can be used for immediate qualitative and as well as for quantitative assessment of patient skin dose. (authors)

  18. SU-E-T-425: Feasibility of Establishing a National Postal Dosimetry System Using EBT3 Gafchromic Film

    International Nuclear Information System (INIS)

    Abdel-Rahman, W; Mohammed, I; Bamajboor, S

    2015-01-01

    Purpose: To evaluate the performance of a postal dosimeter consisting of EBT3 films and a miniature Lucite phantom to be used for photon beams output audits for mega-voltage photon beams across the Saudi Arabia. Methods: Two batches of EBT3 films were calibrated in 5 mega-voltage photon beams with energies ranging between 6 MV and 18 MV in the dose range between 0 to 4 Gy. A 4×4×20 cm 3 lucite phantom was constructed to be used as an irradiation phantom. The lucite irradiation phantom consists of two parts and allows for embedding a 2.5×2.0 cm 2 EBT3 film at 8.9 cm depth. Factors that convert the dose measured in the lucite irradiation phantom at 8.9 cm depth to the dose at 10 cm depth in water for 10×10 cm 2 field for different photon energies were calculated using the dosxyznrc/EGSnrcMP user code. The performance of the proposed postal dosimeter was tested in 17 different photon beams across 4 radiotherapy centres in Saudi Arabia. The outputs of the 17 beams are monitored by either the International Atomic Energy Agency or by the Radiological Physics Centre. Results: For the 17 photon beams, the average of the ratios of measured to stated outputs was 1.01 ± 0.02 and with a maximum ratio of 1.05. Conclusion: The results of our work indicate that the proposed postal dosimetry system can be used for national auditing of outputs for mega-voltage photon beams. The service can be offered to other national radiotherapy centres or to a be used for credentialing of centres participating in national trials

  19. SU-E-T-425: Feasibility of Establishing a National Postal Dosimetry System Using EBT3 Gafchromic Film

    Energy Technology Data Exchange (ETDEWEB)

    Abdel-Rahman, W [KING FAHAD SPECIALIST HOSPITAL, Dammam, Eastern Province (Saudi Arabia); Mohammed, I [King Fahd University of Petroleum & Minerals, Dhahran, Eastern Province (Saudi Arabia); Bamajboor, S [Prince Sultan Military Medical City, Riyadh, Riyadh (Saudi Arabia)

    2015-06-15

    Purpose: To evaluate the performance of a postal dosimeter consisting of EBT3 films and a miniature Lucite phantom to be used for photon beams output audits for mega-voltage photon beams across the Saudi Arabia. Methods: Two batches of EBT3 films were calibrated in 5 mega-voltage photon beams with energies ranging between 6 MV and 18 MV in the dose range between 0 to 4 Gy. A 4×4×20 cm{sup 3} lucite phantom was constructed to be used as an irradiation phantom. The lucite irradiation phantom consists of two parts and allows for embedding a 2.5×2.0 cm{sup 2} EBT3 film at 8.9 cm depth. Factors that convert the dose measured in the lucite irradiation phantom at 8.9 cm depth to the dose at 10 cm depth in water for 10×10 cm{sup 2} field for different photon energies were calculated using the dosxyznrc/EGSnrcMP user code. The performance of the proposed postal dosimeter was tested in 17 different photon beams across 4 radiotherapy centres in Saudi Arabia. The outputs of the 17 beams are monitored by either the International Atomic Energy Agency or by the Radiological Physics Centre. Results: For the 17 photon beams, the average of the ratios of measured to stated outputs was 1.01 ± 0.02 and with a maximum ratio of 1.05. Conclusion: The results of our work indicate that the proposed postal dosimetry system can be used for national auditing of outputs for mega-voltage photon beams. The service can be offered to other national radiotherapy centres or to a be used for credentialing of centres participating in national trials.

  20. Calibration of EBT2 film using a red-channel PDD method in combination with a modified three-channel technique

    Energy Technology Data Exchange (ETDEWEB)

    Chang, Liyun, E-mail: liyunc@isu.edu.tw [Departments of Medical Imaging and Radiological Sciences, I-Shou University, Kaohsiung 82445, Taiwan (China); Ho, Sheng-Yow [Department of Nursing, Chang Jung Christian University, Tainan 73657, Taiwan and Department of Radiation Oncology, Chi Mei Medical Center, Liouying, Tainan 73657, Taiwan (China); Lee, Tsair-Fwu [Medical Physics and Informatics Laboratory, Department of Electronics Engineering, National Kaohsiung University of Applied Sciences, Kaohsiung 80778, Taiwan (China); Yeh, Shyh-An [Department of Medical Imaging and Radiological Sciences, I-Shou University, Kaohsiung 82445, Taiwan and Department of Radiation Oncology, E-Da Hospital, Kaohsiung 82445, Taiwan (China); Ding, Hueisch-Jy [Department of Medical Imaging and Radiological Sciences, I-Shou University, Kaohsiung 82445, Taiwan (China); Chen, Pang-Yu, E-mail: pangyuchen@yahoo.com.tw [Department of Radiation Oncology, Sinlau Christian Hospital, Tainan 70142, Taiwan (China)

    2015-10-15

    Purpose: Ashland Inc. EBT2 and EBT3 films are widely used in quality assurance for radiation therapy; however, there remains a relatively high degree of uncertainty [B. Hartmann, M. Martisikova, and O. Jakel, “Homogeneity of Gafchromic EBT2 film,” Med. Phys. 37, 1753–1756 (2010)]. Micke et al. (2011) recently improved the spatial homogeneity using all color channels of a flatbed scanner; however, van Hoof et al. (2012) pointed out that the corrected nonuniformity still requires further investigation for larger fields. To reduce the calibration errors and the uncertainty, the authors propose a new red-channel percentage-depth-dose method in combination with a modified three-channel technique. Methods: For the ease of comparison, the EBT2 film image used in the authors’ previous study (2012) was reanalyzed using different approaches. Photon beams of 6-MV were delivered to two different films at two different beam on times, resulting in the absorption doses of ranging from approximately 30 to 300 cGy at the vertical midline of the film, which was set to be coincident with the central axis of the beam. The film was tightly sandwiched in a 30{sup 3}-cm{sup 3} polystyrene phantom, and the pixel values for red, green, and blue channels were extracted from 234 points on the central axis of the beam and compared with the corresponding depth doses. The film was first calibrated using the multichannel method proposed by Micke et al. (2010), accounting for nonuniformities in the scanner. After eliminating the scanner and dose-independent nonuniformities, the film was recalibrated via the dose-dependent optical density of the red channel and fitted to a power function. This calibration was verified via comparisons of the dose profiles extracted from the films, where three were exposed to a 60° physical wedge field and three were exposed to composite fields, and all of which were measured in a water phantom. A correction for optical attenuation was implemented, and

  1. Absolute dose measurement Gafchromic R EBT2 movies. Case Study of Kaposis sarcoma

    International Nuclear Information System (INIS)

    Pereira, L.; Moral, F. del; Meilan, E.; Azevedo Gomes, J. C. de; Tejeiro Garcia, A. G.; Andrade Alvarez, B.; Vazquez, J.; Nieto, I.; Medal, D.; Lopez Medina, A.; Francisco, S.; Salgado, M.; Munoz, V.

    2011-01-01

    Because of its high spatial resolution, low energy dependence and good response over a wide energy range, EBT2 Gafchromic films are widely used in many applications in radiotherapy for measuring relative dose. Despite being the most common use can be used to measure absolute dose. This text is an example of using films as EBT2 for in vivo absolute dose in a Kaposis sarcoma.

  2. Verification of an algorithm of cono collapsed through the IAEA TECDOC 1583 protocol and dosimetry with radiochromic films

    International Nuclear Information System (INIS)

    Martin-Viera Cueto, J. A.; Benitez Villegas, E. M.; Bodineau Gil, C.; Parra Osorio, V.; Garcia Pareja, S.; Casado Villalon, F. J.

    2013-01-01

    The objective of this study is to verify the characterization of the collapsed cone algorithm of an SP using this Protocol. In addition, given that it only offers details of dose values measured at discrete points, measures are complemented by a gamma test distributions 2D of doses in different cases using film radiochromic. (Author)

  3. Dose profile variation with voltage in head CT scans using radiochromic films

    Science.gov (United States)

    Mourão, A. P.; Alonso, T. C.; DaSilva, T. A.

    2014-02-01

    The voltage source used in an X-ray tube is an important part of defining the generated beam spectrum energy profile. The X-ray spectrum energy defines the X-ray beam absorption as well as the characteristics of the energy deposition in an irradiated object. Although CT scanners allow one to choose between four different voltage values, most of them employ a voltage of 120 kV in their scanning protocols, regardless of the patient characteristics. Based on this fact, this work investigated the deposited dose in a polymethyl methacrylate (PMMA) cylindrical head phantom. The entire volume was irradiated twice. Two CT scanning protocols were used with two different voltage values: 100 and 120 kV. The phantom volume was irradiated, and radiochromic films were employed to record dose profiles. Measurements were conducted with a calibrated pencil ionization chamber, which was positioned in the center and in four peripheral bores of the head PMMA phantom, to calibrate the radiochromic films. The central slice was then irradiated. This procedure allowed us to find the conversion factors necessary to obtain dose values recorded in the films. The data obtained allowed us to observe the dose variation profile inside the phantom head as well as in the peripheral and central regions. The peripheral region showed higher dose values than those of the central region for scans using both voltage values: approximately 31% higher for scanning with 120 kV and 25% higher with 100 kV. Doses recorded with the highest voltage are significantly higher, approximately 50% higher in the peripheral region and 40% higher in the central region. A longitudinal variation could be observed, and the maximum dose was recorded at the peripheral region, at the midpoint of the longitudinal axis. The obtained results will most likely contribute to the dissemination of proper procedure as well as to optimize dosimetry and tests of quality control in CT because the choice of protocols with different voltage

  4. Fractionated stereotactic radiotherapy: A method to evaluate geometric and dosimetric uncertainties using radiochromic films

    International Nuclear Information System (INIS)

    Coscia, Gianluca; Vaccara, Elena; Corvisiero, Roberta; Cavazzani, Paolo; Ruggieri, Filippo Grillo; Taccini, Gianni

    2009-01-01

    In the authors' hospital, stereotactic radiotherapy treatments are performed with a Varian Clinac 600C equipped with a BrainLAB m3 micro-multileaf-collimator generally using the dynamic conformal arc technique. Patient immobilization during the treatment is achieved with a fixation mask supplied by BrainLAB, made with two reinforced thermoplastic sheets fitting the patient's head. With this work the authors propose a method to evaluate treatment geometric accuracy and, consequently, to determine the amount of the margin to keep in the CTV-PTV expansion during the treatment planning. The reproducibility of the isocenter position was tested by simulating a complete treatment on the anthropomorphic phantom Alderson Rando, inserting in between two phantom slices a high sensitivity Gafchromic EBT film, properly prepared and calibrated, and repeating several treatment sessions, each time removing the fixing mask and replacing the film inside the phantom. The comparison between the dose distributions measured on films and computed by TPS, after a precise image registration procedure performed by a commercial piece of software (FILMQA, 3cognition LLC (Division of ISP), Wayne, NJ), allowed the authors to measure the repositioning errors, obtaining about 0.5 mm in case of central spherical PTV and about 1.5 mm in case of peripheral irregular PTV. Moreover, an evaluation of the errors in the registration procedure was performed, giving negligible values with respect to the quantities to be measured. The above intrinsic two-dimensional estimate of treatment accuracy has to be increased for the error in the third dimension, but the 2 mm margin the authors generally use for the CTV-PTV expansion seems adequate anyway. Using the same EBT films, a dosimetric verification of the treatment planning system was done. Measured dose values are larger or smaller than the nominal ones depending on geometric irradiation conditions, but, in the authors' experimental conditions, always

  5. Dose determination with nitro blue tetrazolium containing radiochromic dye films by measuring absorbed and reflected light

    DEFF Research Database (Denmark)

    Kovács, A.; Baranyai, M.; Wojnárovits, L.

    2000-01-01

    determination in a wide dose range both by absorbance and reflectance measurements. The concept of measuring reflected light from dose labels has been discussed earlier and emerged recently due to the requirement of introducing semiquantitative label dose indicators for quarantine control. The usefulness...... of the method was studied using the newly developed radiochromic dye films as well as already existing ones. (C) 2000 Elsevier Science Ltd. All rights reserved....

  6. Practical use of Gafchromic(®) EBT films in electron beams for in-phantom dose distribution measurements and monitor units verification.

    Science.gov (United States)

    El Barouky, Jad; Fournier-Bidoz, Nathalie; Mazal, Alejandro; Fares, Georges; Rosenwald, Jean-Claude

    2011-04-01

    The possibility of using the Gafchromic(®) EBT films parallel to incident electron beams was assessed in order to facilitate quality assurance tests for electron dose calculation algorithms. Calibration curves were made for electron energies of 6, 9 and 12MeV. A set-up was suggested for EBT film irradiation parallel to the beam, and the dose measurements were compared to Ionization Chamber (IC) measurements in standard and small electrons beams. A more complex Quality Assurance (QA) set-up was performed with the cylindrical CARPET(®) phantom in order to test our Treatment Planning System (TPS) (Eclipse, Varian Medical Systems, Palo Alto, California) for the clinical situation of a chest wall electron beam therapy. Two dimensional dose distribution and gamma index results were compared to the calculated distribution given by the TPS. The reproducibility was found to be better than 1.5%. We found that applying strong pressure and aligning carefully the film edge with the phantom surface, as recommended for radiographic films, did not completely eliminate the air gap effect. Adding an ultrasound transmission gel and 2 complementary EBT films on the surface gave satisfactory results. The absolute dose for the reference 10×10cm(2) field was always within 1% of IC measurements and for smaller elongated fields (5×10, 4×10 and 3×10cm(2)) the mean difference was -1.4% for the three energies. The mean difference with the IC measurements in R(100), R(90) and R(50) was 0.9mm for all fields and for the three energies. The mean difference in the width of the 90% and the 50% isodoses at R(100) was 0.6mm. With the CARPET(®) phantom set-up very good agreement was found in the 2D dose distribution; 99% of the points satisfied the γdose distribution if ultrasound gel and overlying perpendicular films are added on the phantom surface. Copyright © 2010 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  7. Measurements of spatial distribution of absorbed dose in proton therapy with Gafchromic EBT3

    International Nuclear Information System (INIS)

    Gambarini, G.; Regazzoni, V.; Grisotto, S.; Artuso, E.; Giove, D.; Borroni, M.; Carrara, M.; Pignoli, E.; Mirandola, A.; Ciocca, M.

    2014-08-01

    A study of the response of EBT3 films has been carried out. Light transmittance images (around 630 nm) were acquired by means of a Ccd camera. The difference of optical density was assumed as dosimeter response. Calibration was performed by means of 60 Co photons, at a radiotherapy facility. A study of the response variation during the time after exposure has been carried out. EBT3 films were exposed, in a solid-water phantom, to proton beams of various energies and the obtained depth-dose profiles were compared with those measured with a ionization chamber. As expected, in the Bragg peak region the values obtained with EBT3 films were lower than those obtained with the ionization chamber. The ratio of such values was evaluated, along dose profiles, for each utilized energy. A method for correcting the data measured with EBT3 has been proposed and tested. The results confirm that the method can be advantageously applied for obtaining spatial distribution of the absorbed dose in proton therapy. (author)

  8. Measurements of spatial distribution of absorbed dose in proton therapy with Gafchromic EBT3

    Energy Technology Data Exchange (ETDEWEB)

    Gambarini, G.; Regazzoni, V.; Grisotto, S.; Artuso, E.; Giove, D. [Universita degli Studi di Milano, Department of Physics, via Celoria 16, 20133 Milano (Italy); Borroni, M.; Carrara, M.; Pignoli, E. [Fondazione IRCCS, Istituto Nazionale dei Tumori di Milano, Medical Physics Unit, via Giacomo Venezian 16, 20133 Milano (Italy); Mirandola, A.; Ciocca, M., E-mail: grazia.gambarini@mi.infn.it [Centro Nazionale Adroterapia Oncologica, Medical Physics Unit, Strada Campeggi 53, 27100 Pavia (Italy)

    2014-08-15

    A study of the response of EBT3 films has been carried out. Light transmittance images (around 630 nm) were acquired by means of a Ccd camera. The difference of optical density was assumed as dosimeter response. Calibration was performed by means of {sup 60}Co photons, at a radiotherapy facility. A study of the response variation during the time after exposure has been carried out. EBT3 films were exposed, in a solid-water phantom, to proton beams of various energies and the obtained depth-dose profiles were compared with those measured with a ionization chamber. As expected, in the Bragg peak region the values obtained with EBT3 films were lower than those obtained with the ionization chamber. The ratio of such values was evaluated, along dose profiles, for each utilized energy. A method for correcting the data measured with EBT3 has been proposed and tested. The results confirm that the method can be advantageously applied for obtaining spatial distribution of the absorbed dose in proton therapy. (author)

  9. Assessment of the doses received by patients with pacemakers who have received treatment of external radiotherapy; Evaluacion de la dosis recibida por pacientes con marcapasos que han recibido tratamiento de radioterapia externa

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez Alarcon, C.; Carrasco Herrera, M. A.; Vicent Grandado, D.; Gutierrez Ramos, S. M.; Herrador Cordoba, M.

    2013-07-01

    The objective of this study is to estimate with system planning, TPS the dose that would receive the pacemaker and its comparison with the dose received during the treatment a film radiochromic EBT2 located in the area where it is located. (Author)

  10. Numerical modeling of the EBT-S ion-cyclotron heating experiment

    International Nuclear Information System (INIS)

    Sperling, J.L.; Hamasaki, S.; Klein, H.H.; Krall, N.A.

    1980-01-01

    To determine the effect of ion-cyclotron heating on EBT-S plasma parameters, a one-dimensional, time dependent neoclassical model of plasma particle and energy transport was used. For EBT-S the code was run with the following parameters: B/sub O/ = 0.7 tesla (axial field at the midplane), B/sub O/ = 1.4 tesla (axial field at the throat), R/sub T/ = 150 cm (major radius), a = 15 cm

  11. Evaluation of dose calculation algorithms for the electron beams used in radiotherapy. Comparison with radiochromic film measurements

    International Nuclear Information System (INIS)

    El Barouky, Jad

    2011-01-01

    In radiotherapy, the dose calculation accuracy is crucial for the quality and the outcome of the treatments. The purpose of our study was to evaluate the accuracy of dose calculation algorithms for electron beams in situations close to clinical conditions. A new practical approach of radiochromic film dosimetry was developed and validated especially for difficult situations. An accuracy of 3.1% and 2.6% was achieved for absolute and relative dosimetry respectively. Using this technique a measured database of dose distributions was developed to form the basis of several fast and efficient Quality Assurance tests. Such tests are intended to be used also when the dose calculation algorithm is changed or the Treatment Planning System replaced. Pencil Beam and Monte Carlo dose calculations were compared to the measured data for simple geometrical phantom setups. They both gave similar results for obliquity, surface irregularity and extended SSD tests but the Monte Carlo calculation was more accurate in presence of heterogeneities. The same radiochromic film dosimetry method was applied to film cuts inserted into anthropomorphic phantoms providing a 2D dose distribution for any transverse plan. This allowed us to develop clinical test that can be also used for internal Quality Assurance purposes. As for simpler geometries, the Monte Carlo calculations showed better agreement with the measured data than the Pencil Beam calculation, especially in presence of heterogeneities such as lungs, cavities and bones. (author) [fr

  12. Fractionated stereotactic radiotherapy: A method to evaluate geometric and dosimetric uncertainties using radiochromic films

    Energy Technology Data Exchange (ETDEWEB)

    Coscia, Gianluca; Vaccara, Elena; Corvisiero, Roberta; Cavazzani, Paolo; Ruggieri, Filippo Grillo; Taccini, Gianni [S. C. Fisica Sanitaria, E.O. Ospedali Galliera di Genova, Via Alessandro Volta, 8 16128 Genova (Italy); S. C. Radioterapia, E.O. Ospedali Galliera di Genova, Via Alessandro Volta, 8 16128 Genova (Italy); S. C. Fisica Sanitaria, E.O. Ospedali Galliera di Genova, Via Alessandro Volta, 8 16128 Genova (Italy)

    2009-07-15

    In the authors' hospital, stereotactic radiotherapy treatments are performed with a Varian Clinac 600C equipped with a BrainLAB m3 micro-multileaf-collimator generally using the dynamic conformal arc technique. Patient immobilization during the treatment is achieved with a fixation mask supplied by BrainLAB, made with two reinforced thermoplastic sheets fitting the patient's head. With this work the authors propose a method to evaluate treatment geometric accuracy and, consequently, to determine the amount of the margin to keep in the CTV-PTV expansion during the treatment planning. The reproducibility of the isocenter position was tested by simulating a complete treatment on the anthropomorphic phantom Alderson Rando, inserting in between two phantom slices a high sensitivity Gafchromic EBT film, properly prepared and calibrated, and repeating several treatment sessions, each time removing the fixing mask and replacing the film inside the phantom. The comparison between the dose distributions measured on films and computed by TPS, after a precise image registration procedure performed by a commercial piece of software (FILMQA, 3cognition LLC (Division of ISP), Wayne, NJ), allowed the authors to measure the repositioning errors, obtaining about 0.5 mm in case of central spherical PTV and about 1.5 mm in case of peripheral irregular PTV. Moreover, an evaluation of the errors in the registration procedure was performed, giving negligible values with respect to the quantities to be measured. The above intrinsic two-dimensional estimate of treatment accuracy has to be increased for the error in the third dimension, but the 2 mm margin the authors generally use for the CTV-PTV expansion seems adequate anyway. Using the same EBT films, a dosimetric verification of the treatment planning system was done. Measured dose values are larger or smaller than the nominal ones depending on geometric irradiation conditions, but, in the authors' experimental

  13. Quality control of specific patient in radiotherapy with modulated intensity

    International Nuclear Information System (INIS)

    Aberbuj, P D; Tapia Coca, R C

    2012-01-01

    In this work we comment the details of the patient specific quality controls of the first Intensity Modulated Radiotherapy treatment done at Roffo Institute. These controls consisted in two sets of measurements: absolute dose with ionization chamber and relative dose with two dosimetric systems (Gafchromic EBT2 radiochromic films and the PTW 729 ionization chambers array). Two of the filters did not pass the dosimetrical tests, and they were manufactured again. The new filters passed the tests. For the relative two-dimensional measurements the radiochromic films had a better performance than the array due to their higher spatial resolution (author)

  14. Measurement of off-axis and peripheral skin dose using radiochromic film

    International Nuclear Information System (INIS)

    Butson, Martin J.; Yu, P.K.N.; Metcalfe, P.E.

    1998-01-01

    A radiotherapy skin dose profile can be obtained with radiochromic film. The central axis skin dose relative to D max for a 10x10cm 2 field size was found to be 22%, 17% and 15.5% for 6 MV, 10 MV and 18 MV photon beams. Peripheral dose increased with increasing field size. At 10 MV the skin dose 2 cm outside the geometric field edge was measured as 6%, 10% and 17% for 10x10cm 2 , 20x20cm 2 and 30x30cm 2 field sizes respectively. Off-axis skin dose decreased as distance increased from central axis for fields with Perspex block trays. For a 20x20cm 2 field, an approximately 5-8% drop in percentage skin dose was observed from central axis to the beam edge. (author)

  15. Temperature, humidity and time. Combined effects on radiochromic film dosimeters

    DEFF Research Database (Denmark)

    Abdel-Fattah, A.A.; Miller, A.

    1996-01-01

    The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Riso B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60 degrees C for irradiation by Co-60 photons and 10-Me......V electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is +0.......25 +/- 0.1% per degrees C for the FWT-60-00 dosimeters and +0.5 +/- 0.1% per degrees C For Riso B3 dosimeters at temperatures between 20 and 50 degrees C and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger...

  16. Dosimetric reproduction of a left-breast 3DCRT field-in-field radiation therapy planning in an anthropomorphic and anthropometric phantom

    Energy Technology Data Exchange (ETDEWEB)

    Nogueira, Luciana B., E-mail: lucibn19@yahoo.com.br, E-mail: jonymarques@uol.com.br [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Anatomia e Imagem; Barsanelli, Cristiane; Geraldo, Jony M., E-mail: cbarsanelli@yahoo.com.br [Hospital Luxemburgo, Instituto Mário Penna, Belo Horizonte, MG (Brazil); Aquino, Jean Carlos; Campos, Tarcísio P. Ribeiro, E-mail: jeancarlosaquino@outlook.com, E-mail: tprcampos@yahoo.com.br [Universidade Federal de Minas Gerais (UGMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear

    2017-07-01

    The proposal of this study was to reproduce the dosimetry established in a treatment planning system (TPS) following a 3D conformational radiation therapy (3DCRT) protocol of two parallel-opposite fields applied to the left-breast in a thorax phantom, with the use of the field-in-field technique. Computed tomography (CT) images of the anthropomorphic and anthropometric phantom of the thorax with mobile breasts were generated. This phantom was developed by the NRI / UFMG research group. The generated images were transferred to the planning system XiO version-5 for the elaboration of the breast tele therapeutic planning with 2 Gy per fraction, in 25 fractions, with prescribed dose of 50 Gy. A set of ten EBT2 radiochromic films were irradiated at different doses. The values of RGB (Red, Green, Blue) of the radiochromic films were obtained by scanning and data transformed in optical density (OD), whose values were used to construct the calibration curve. EBT2 radiochromic films were positioned outside and inside of the thorax phantom: internally in the right and left lungs, on the face of the heart, in the glandular breast tissue-equivalent (TE) and in the left breast skin. After phantom radiation at the linear accelerator 6 MV Elekta Precise reproducing the 3DCRT, the radiochromic films were digitized after 24 h of exposure. The measurements of the intensities of the films in RGB were measured in the software ImageJ, transformed in optical density and converted in bidimensional dose distributions, applying the calibration curve. The experimental dosimetric data were analyzed and compared with values generated in the TPS. In addition, graphics and dose-volume histograms (DVH) were developed. The dose measurements in the glandular-TE in breast did not present statistically significant differences in relation to values at equivalent positions generated in the TPS. The organs at risk received doses below the reference values, according to TPS. It was verified the

  17. Dosimetric reproduction of a left-breast 3DCRT field-in-field radiation therapy planning in an anthropomorphic and anthropometric phantom

    International Nuclear Information System (INIS)

    Nogueira, Luciana B.; Aquino, Jean Carlos; Campos, Tarcísio P. Ribeiro

    2017-01-01

    The proposal of this study was to reproduce the dosimetry established in a treatment planning system (TPS) following a 3D conformational radiation therapy (3DCRT) protocol of two parallel-opposite fields applied to the left-breast in a thorax phantom, with the use of the field-in-field technique. Computed tomography (CT) images of the anthropomorphic and anthropometric phantom of the thorax with mobile breasts were generated. This phantom was developed by the NRI / UFMG research group. The generated images were transferred to the planning system XiO version-5 for the elaboration of the breast tele therapeutic planning with 2 Gy per fraction, in 25 fractions, with prescribed dose of 50 Gy. A set of ten EBT2 radiochromic films were irradiated at different doses. The values of RGB (Red, Green, Blue) of the radiochromic films were obtained by scanning and data transformed in optical density (OD), whose values were used to construct the calibration curve. EBT2 radiochromic films were positioned outside and inside of the thorax phantom: internally in the right and left lungs, on the face of the heart, in the glandular breast tissue-equivalent (TE) and in the left breast skin. After phantom radiation at the linear accelerator 6 MV Elekta Precise reproducing the 3DCRT, the radiochromic films were digitized after 24 h of exposure. The measurements of the intensities of the films in RGB were measured in the software ImageJ, transformed in optical density and converted in bidimensional dose distributions, applying the calibration curve. The experimental dosimetric data were analyzed and compared with values generated in the TPS. In addition, graphics and dose-volume histograms (DVH) were developed. The dose measurements in the glandular-TE in breast did not present statistically significant differences in relation to values at equivalent positions generated in the TPS. The organs at risk received doses below the reference values, according to TPS. It was verified the

  18. Characterization of Thymol blue Radiochromic dosimeters for high dose applications

    Directory of Open Access Journals (Sweden)

    Feras M. Aldweri

    2018-03-01

    Full Text Available Thymol blue (TB solutions and Thymol blue Polyvinyl Alcohol (TB-PVA films have been introduced as Radiochromic dosimeter for high dose applications. The dosimeters were irradiated with gamma ray (60Co source from 5 to 30 kGy for film, and from 0.150 kGy to 4 kGy for solution. The optical density of unirradiated and irradiated TB solution as well as TB-PVA film dosimeters were studied in terms of absorbance at 434 nm using UV/VIS spectrophotometer. The effects of scan temperature, light pre-gamma irradiation, dose rate, relative humidity and stability of the absorbance of solutions and films after irradiation were investigated. We found the dose sensitivity of TB solution and TB-PVA film dosimeters increases significantly with increases of the absorbed dose as well as with the increases of TB dye concentrations. The useful dose range of developed TB solutions and TB-PVA films dosimeters is in the range 0.125–1 kGy and of 5–20 kGy, respectively. Keywords: Dose sensitivity, Radio-chromic dosimeter, Thymol blue, Absorbance, Concentrations

  19. Verification of absorbed dose rates in reference beta radiation fields: measurements with an extrapolation chamber and radiochromic film

    International Nuclear Information System (INIS)

    Reynaldo, S. R.; Benavente C, J. A.; Da Silva, T. A.

    2015-10-01

    Beta Secondary Standard 2 (Bss 2) provides beta radiation fields with certified values of absorbed dose to tissue and the derived operational radiation protection quantities. As part of the quality assurance, metrology laboratories are required to verify the reliability of the Bss-2 system by performing additional verification measurements. In the CDTN Calibration Laboratory, the absorbed dose rates and their angular variation in the 90 Sr/ 90 Y and 85 Kr beta radiation fields were studied. Measurements were done with a 23392 model PTW extrapolation chamber and with Gafchromic radiochromic films on a PMMA slab phantom. In comparison to the certificate values provided by the Bss-2, absorbed dose rates measured with the extrapolation chamber differed from -1.4 to 2.9% for the 90 Sr/ 90 Y and -0.3% for the 85 Kr fields; their angular variation showed differences lower than 2% for incidence angles up to 40-degrees and it reached 11% for higher angles, when compared to ISO values. Measurements with the radiochromic film showed an asymmetry of the radiation field that is caused by a misalignment. Differences between the angular variations of absorbed dose rates determined by both dosimetry systems suggested that some correction factors for the extrapolation chamber that were not considered should be determined. (Author)

  20. Electron-confinement studies on EBT-S using soft-x-ray techniques

    International Nuclear Information System (INIS)

    Hillis, D.L.; Haste, G.R.; Berry, L.A.

    1982-08-01

    Soft x-ray bremsstrahlung measurements have been performed on the ELMO Bumpy Torus (EBT-S) plasma to determine the electron temperature T/sub e/ and electron density n/sub e/ using a calibrated Si(Li) detector over a wide range of operating conditions. The purpose of this paper is to outline the necessary assumptions and essential x-ray techniques that are inherent in soft x-ray measurements in order to investigate the electron heating and confinement properties of EBT-S. In addition, by utilizing the electron density as determined by the soft x-ray measurements, the previous EBT-S confinement analyses have been extended. The steady-state plasma of EBT-S is heated by microwaves using a cw gyrotron that can operate up to power levels of 200 kW. From the soft x-ray measurements, both the electron temperature and density are found to increase at higher microwave power levels. For operation at microwave power levels of 200 kW, T/sub e/ approaches 1 keV while n/sub e/ approaches 1.2 x 10 12 cm -3 . In general, confinement properties are found to improve with increased microwave power. The data are compared with neoclassical transport scaling and the electron transport is found to be collisionless (nu/Ω < 1) as well as neoclassical

  1. Computed tomography scan optimization using head phantom and radiochromic films

    Energy Technology Data Exchange (ETDEWEB)

    Fernandes, Lorena C.; Santana, P.C.; Velasquez, Carlos E.; Mourao, Arnaldo P., E-mail: lorena.cfernandes@hotmail.com, E-mail: carlosvelcab@hotmail.com, E-mail: pridili@gmail.com, E-mail: apmourao@gmail.com [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte (Brazil). Departamento de Engenharia Nuclear; Centro de Imagem Molecular do INCT-Medicina Molecular (CIMol/UFMG) Belo Horizonte (Brazil); Centro Federal de Educação Tecnológica de Minas Gerais (CEFET-MG), Belo Horizonte (Brazil)

    2017-11-01

    The research and development in technology applied to computed tomography has been improve in the image quality, resulting in the best identification of diseases, and therefore an increase in the number of exams, among them the head exams can be highlighted. In order to promote a radioprotection and reduction on the dose of the general public some international agencies have stipulated dose limits to be followed and the implementation of the principles of ALARA in all establishments that use ionizing radiation. One of these principles is the optimization that should be treated with substantial attention, because the reduction of radiation doses can be feasible as long as it does not compromise the tomographic images; such practice is difficult to perform due to the lack of proper guidance. In optimization of the CT exams, not only the lowest dose is evaluated to obtain diagnostic image, but also should be knowledge the dose distribution throughout the scanned area. In this work were used a cylindrical head phantom of PMMA, a GE Discovery CT scan with 64 channels, and radiochromic films. The films were positioned in the phantom in their central region for the dose evaluation using the automatic control for the voltages of 80, 100 and 120 kV. The images were acquired from the scan of the phantom and the film readings were obtained through digital images. The results show an evaluation of the longitudinal kerma profiles, dose delivered, and the image noise was also observed using the central slice images. (author)

  2. Computed tomography scan optimization using head phantom and radiochromic films

    International Nuclear Information System (INIS)

    Fernandes, Lorena C.; Santana, P.C.; Velasquez, Carlos E.; Mourao, Arnaldo P.

    2017-01-01

    The research and development in technology applied to computed tomography has been improve in the image quality, resulting in the best identification of diseases, and therefore an increase in the number of exams, among them the head exams can be highlighted. In order to promote a radioprotection and reduction on the dose of the general public some international agencies have stipulated dose limits to be followed and the implementation of the principles of ALARA in all establishments that use ionizing radiation. One of these principles is the optimization that should be treated with substantial attention, because the reduction of radiation doses can be feasible as long as it does not compromise the tomographic images; such practice is difficult to perform due to the lack of proper guidance. In optimization of the CT exams, not only the lowest dose is evaluated to obtain diagnostic image, but also should be knowledge the dose distribution throughout the scanned area. In this work were used a cylindrical head phantom of PMMA, a GE Discovery CT scan with 64 channels, and radiochromic films. The films were positioned in the phantom in their central region for the dose evaluation using the automatic control for the voltages of 80, 100 and 120 kV. The images were acquired from the scan of the phantom and the film readings were obtained through digital images. The results show an evaluation of the longitudinal kerma profiles, dose delivered, and the image noise was also observed using the central slice images. (author)

  3. Validation of the entire 2D array Octavius by radiochromic films; Validacion del conjunto Octavius 2D array mediante peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Iriondo Igerabide, U.; Former Forner, A.; Otal Palacin, A.; Martin Albina, M. L.; Lozares Cordero, S.; Maneru Camara, F.; Pellejero Pellejero, S.; Miquelez Alonso, S.; Soto Prados, P. M.; Rubio Arroniz, A.

    2011-07-01

    We have validated the 2D-array Octavius set for verification of IMRT plans complete, radiating with the same angles of incidence in the patient's actual plan. This has been taken as reference measurements with radiochromic films, since they are almost isotropic response, higher resolution and we already have experience with them.

  4. I-125 ROPES eye plaque dosimetry: validation of a commercial 3D ophthalmic brachytherapy treatment planning system and independent dose calculation software with GafChromic® EBT3 films.

    Science.gov (United States)

    Poder, Joel; Corde, Stéphanie

    2013-12-01

    The purpose of this study was to measure the dose distributions for different Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaques loaded with I-125 (model 6711) seeds using GafChromic(®) EBT3 films, in order to verify the dose distributions in the Plaque Simulator™ (PS) ophthalmic 3D treatment planning system. The brachytherapy module of RADCALC(®) was used to independently check the dose distributions calculated by PS. Correction factors were derived from the measured data to be used in PS to account for the effect of the stainless steel ROPES plaque backing on the 3D dose distribution. Using GafChromic(®) EBT3 films inserted in a specially designed Solid Water™ eye ball phantom, dose distributions were measured three-dimensionally both along and perpendicular to I-125 (model 6711) loaded ROPES eye plaque's central axis (CAX) with 2 mm depth increments. Each measurement was performed in full scatter conditions both with and without the stainless steel plaque backing attached to the eye plaque, to assess its effect on the dose distributions. Results were compared to the dose distributions calculated by Plaque Simulator™ and checked independently with RADCALC(®). The EBT3 film measurements without the stainless steel backing were found to agree with PS and RADCALC(®) to within 2% and 4%, respectively, on the plaque CAX. Also, RADCALC(®) was found to agree with PS to within 2%. The CAX depth doses measured using EBT3 film with the stainless steel backing were observed to result in a 4% decrease relative to when the backing was not present. Within experimental uncertainty, the 4% decrease was found to be constant with depth and independent of plaque size. Using a constant dose correction factor of T = 0.96 in PS, where the calculated dose for the full water scattering medium is reduced by 4% in every voxel in the dose grid, the effect of the plaque backing was accurately modeled in the planning system. Off-axis profiles

  5. Quality assurance in RapidArc with Alderson anthropomorphic phantom using radiochromic film in comparison to MATLAB; Controle de qualidade em RapidArc com simulador de corpo humano antropomorfico Alderson utilizando filme radiocromico em comparacao ao MATLAB

    Energy Technology Data Exchange (ETDEWEB)

    Garcia, Paulo L.; Silva, Leonardo P.; Santos, Maira R.; Trindade, Cassia; Martins, Lais P.; Batista, Delano V.S., E-mail: Paulo8_lgarcia@hotmail.com [Instituto Nacional de Cancer (INCA), Rio de Janeiro, RJ (Brazil); Alves, Victor G. [Instituto Nacional de Cancer (SQRIS/INCA), Rio de Janeiro, RJ (Brazil). Servico de Qualidade em Radiacoes Ionizantes

    2012-12-15

    This paper presented the quality control for RapidArc using an Alderson human body phantom and radiochromic film as an alternative system to approve the treatment plan for brain tumor. Thus, it was comprised the dose distributions provided by the treatment planning system with those measured by the film radiochromic. The gamma index (Γ) analysis, to verify the acceptability of the dose distribution, was 95% of approved points, with the mostly non-compliance points in regions near the PTV’s edges. These non-compliance points may be associated to transmission blades aspects, because the regions near the edges present significant losses compared to the central areas. Also, MATLAB has proved an effective tool for that measurements and it can be used in quality assurance programs. (author)

  6. SU-E-T-746: The Use of Radiochromic Film Analyzed with Three Channel Dosimetry as a Secondary Patient-Specific QA Tool for Small SBRT Fields

    International Nuclear Information System (INIS)

    Hadsell, M; Holcombe, C; Chin, E; Hsu, A

    2015-01-01

    Introduction: As diagnostic techniques become more sensitive and targeting methods grow in accuracy, target volumes continue to shrink and SBRT becomes more prevalent. Due to this fact, patient-specific QA must also enhance resolution and accuracy in order to verify dose delivery in these volumes. It has been suggested that when measuring small fields at least two separate detectors be used to verify delivered dose. Therefore, we have instituted a secondary patient QA verification for small (<3cm) SBRT fields using Gafchromic EBT2 film. Methods: Films were cross-calibrated using a Farmer chamber in plastic water at reference conditions as defined by TG-51. Films were scanned, and an RGB calibration curve was created according to best practices published by Ashland, Inc. Four SBRT cases were evaluated both with the Scandidos Delta4 and with EBT2 films sandwiched in plastic water. Raw values obtained from the film were converted to dose using an in-house algorithm employing all three color channels to increase accuracy and dosimetric range. Gamma and dose profile comparisons to Eclipse dose calculations were obtained using RIT and compared to values obtained with the Delta4. Results: Film gamma pass rates at 2% and 2mm were similar to those obtained with the Delta4. However, dose difference histograms showed better absolute dose agreement, with the average mean film dose agreeing with calculation to 0.3% and the Delta4 only agreeing to 3.1% across the cases. Additionally, films provided more resolution than the Delta4 and thus their dose profiles better succeeded in diagnosing dose calculation inaccuracies. Conclusion: We believe that the implementation of secondary patient QA using EBT2 film analyzed with all three color channels is an invaluable tool for evaluation of small SBRT fields. Furthermore, we have shown that this method can sometimes provide a more detailed and faithful reproduction of plan dose than the Delta4

  7. A radiochromic film based on leucomalachite green for high-dose dosimetry applications

    International Nuclear Information System (INIS)

    Soliman, Y.S.; Basfar, A.A.; Msalam, R.I.

    2014-01-01

    A colorless polyvinyl butyral film (PVB) based on radiation-sensitive dye of leucomalachite green (LMG) was investigated as a high-dose dosimeter for gamma radiation processing applications in the dose range of 3–150 kGy. The useful applications for such dose range are food irradiation treatment, medical devices sterilization and polymer modification. Gamma irradiation of the film induces a significant intensity of green color, which can be characterized by a main absorption band at 627 nm and a small band at 425 nm. The variation in response of irradiated film stored in the dark and under laboratory light illumination was less than 3% during the first 6 days of storage. The response of film during irradiation was slightly influenced by relative humidity in the range of 12–76%; however, it was significantly affected by temperature in the range of 5–40 °C. The radiation chemical yield was reported to be 6.76 × 10 −6  mol/J at the absorbed dose of 30 kGy for the film containing 6.5% of LMG dye. The overall uncertainty associated with routine dose monitoring would be less than 6% at a 95% confidence level if the dosimeter was being corrected for irradiation conditions and being calibrated with reference standard dosimeter in the production facility. - Highlights: • Development of a radiochromic film based on leucomalachite green dye for radiation processing dosimetry. • The dosimeter useful dose range is 3–150 kGy. • The dosimeter was slightly influenced by humidity levels during irradiation over the range of 12–76%. • The films stored in the dark have a good shelf life with a good stable response after irradiation. • Overall uncertainty of the dosimeter was less than 4.3% at σ

  8. Analysis of the EBT3 Gafchromic film irradiated with 6 MV photons and 6 MeV electrons using reflective mode scanners.

    Science.gov (United States)

    Farah, Nicolas; Francis, Ziad; Abboud, Marie

    2014-09-01

    We explore in our study the effects of electrons and X-rays irradiations on the newest version of the Gafchromic EBT3 film. Experiments are performed using the Varian "TrueBeam 1.6" medical accelerator delivering 6 MV X-ray photons and 6 MeV electron beams as desired. The main interest is to compare the responses of EBT3 films exposed to two separate beams of electrons and photons, for radiation doses ranging up to 500 cGy. The analysis is done on a flatbed EPSON 10000 XL scanner and cross checked on a HP Scanjet 4850 scanner. Both scanners are used in reflection mode taking into account landscape and portrait scanning positions. After thorough verifications, the reflective scanning method can be used on EBT3 as an economic alternative to the transmission method which was also one of the goals of this study. A comparison is also done between single scan configuration including all samples in a single A4 (HP) or A3 (EPSON) format area and multiple scan procedure where each sample is scanned separately on its own. The images analyses are done using the ImageJ software. Results show significant influence of the scanning configuration but no significant differences between electron and photon irradiations for both single and multiple scan configurations. In conclusion, the film provides a reliable relative dose measurement method for electrons and photons irradiations in the medical field applications. Copyright © 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  9. In vivo dosimetry in intraoperative electron radiotherapy: microMOSFETs, radiochromic films and a general-purpose linac.

    Science.gov (United States)

    López-Tarjuelo, Juan; Bouché-Babiloni, Ana; Morillo-Macías, Virginia; de Marco-Blancas, Noelia; Santos-Serra, Agustín; Quirós-Higueras, Juan David; Ferrer-Albiach, Carlos

    2014-10-01

    In vivo dosimetry is desirable for the verification, recording, and eventual correction of treatment in intraoperative electron radiotherapy (IOERT). Our aim is to share our experience of metal oxide semiconductor field-effect transistors (MOSFETs) and radiochromic films with patients undergoing IOERT using a general-purpose linac. We used MOSFETs inserted into sterile bronchus catheters and radiochromic films that were cut, digitized, and sterilized by means of gas plasma. In all, 59 measurements were taken from 27 patients involving 15 primary tumors (seven breast and eight non-breast tumors) and 12 relapses. Data were subjected to an outliers' analysis and classified according to their compatibility with the relevant doses. Associations were sought regarding the type of detector, breast and non-breast irradiation, and the radiation oncologist's assessment of the difficulty of detector placement. At the same time, 19 measurements were carried out at the tumor bed with both detectors. MOSFET measurements ([Formula: see text]  = 93.5 %, sD  =  6.5 %) were not significantly shifted from film measurements ([Formula: see text]  =  96.0 %, sD  =  5.5 %; p  =  0.109), and no associations were found (p = 0.526, p = 0.295,  and p = 0.501, respectively). As regards measurements performed at the tumor bed with both detectors, MOSFET measurements ([Formula: see text]  =  95.0 %, sD  =  5.4 % were not significantly shifted from film measurements ([Formula: see text]  =  96.4 %, sD  =  5.0 %; p  =  0.363). In vivo dosimetry can produce satisfactory results at every studied location with a general-purpose linac. Detector choice should depend on user factors, not on the detector performance itself. Surgical team collaboration is crucial to success.

  10. I-125 ROPES eye plaque dosimetry: Validation of a commercial 3D ophthalmic brachytherapy treatment planning system and independent dose calculation software with GafChromic{sup ®} EBT3 films

    Energy Technology Data Exchange (ETDEWEB)

    Poder, Joel; Corde, Stéphanie [Department of Radiation Oncology, Prince of Wales Hospital, Randwick, NSW 2031 (Australia)

    2013-12-15

    Purpose: The purpose of this study was to measure the dose distributions for different Radiation Oncology Physics and Engineering Services, Australia (ROPES) type eye plaques loaded with I-125 (model 6711) seeds using GafChromic{sup ®} EBT3 films, in order to verify the dose distributions in the Plaque Simulator™ (PS) ophthalmic 3D treatment planning system. The brachytherapy module of RADCALC{sup ®} was used to independently check the dose distributions calculated by PS. Correction factors were derived from the measured data to be used in PS to account for the effect of the stainless steel ROPES plaque backing on the 3D dose distribution.Methods: Using GafChromic{sup ®} EBT3 films inserted in a specially designed Solid Water™ eye ball phantom, dose distributions were measured three-dimensionally both along and perpendicular to I-125 (model 6711) loaded ROPES eye plaque's central axis (CAX) with 2 mm depth increments. Each measurement was performed in full scatter conditions both with and without the stainless steel plaque backing attached to the eye plaque, to assess its effect on the dose distributions. Results were compared to the dose distributions calculated by Plaque Simulator™ and checked independently with RADCALC{sup ®}.Results: The EBT3 film measurements without the stainless steel backing were found to agree with PS and RADCALC{sup ®} to within 2% and 4%, respectively, on the plaque CAX. Also, RADCALC{sup ®} was found to agree with PS to within 2%. The CAX depth doses measured using EBT3 film with the stainless steel backing were observed to result in a 4% decrease relative to when the backing was not present. Within experimental uncertainty, the 4% decrease was found to be constant with depth and independent of plaque size. Using a constant dose correction factor of T= 0.96 in PS, where the calculated dose for the full water scattering medium is reduced by 4% in every voxel in the dose grid, the effect of the plaque backing was

  11. A definitive method for dosimetry with radiochromic movie

    International Nuclear Information System (INIS)

    Miras del Rio, H.; Arrans Lara, R.

    2013-01-01

    Despite the undoubted benefits of radiochromic film as a two-dimensional radiation detector, the manufacturing process and characteristics of scanners make to present some inherent difficulties that result that its use is complex, tedious, and if not performed so appropriate, imprecise. A protocol that simplifies the calibration process while reducing the number of films used for this purpose is proposed systematically corrects several difficulties, both inherent in the manufacture of the film as the reader and, what is more important achieves levels comparable to those of other more widespread use detectors accuracy. (Author)

  12. Verification of absorbed dose rates in reference beta radiation fields: measurements with an extrapolation chamber and radiochromic film

    Energy Technology Data Exchange (ETDEWEB)

    Reynaldo, S. R. [Development Centre of Nuclear Technology, Posgraduate Course in Science and Technology of Radiations, Minerals and Materials / CNEN, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil); Benavente C, J. A.; Da Silva, T. A., E-mail: sirr@cdtn.br [Development Centre of Nuclear Technology / CNEN, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil)

    2015-10-15

    Beta Secondary Standard 2 (Bss 2) provides beta radiation fields with certified values of absorbed dose to tissue and the derived operational radiation protection quantities. As part of the quality assurance, metrology laboratories are required to verify the reliability of the Bss-2 system by performing additional verification measurements. In the CDTN Calibration Laboratory, the absorbed dose rates and their angular variation in the {sup 90}Sr/{sup 90}Y and {sup 85}Kr beta radiation fields were studied. Measurements were done with a 23392 model PTW extrapolation chamber and with Gafchromic radiochromic films on a PMMA slab phantom. In comparison to the certificate values provided by the Bss-2, absorbed dose rates measured with the extrapolation chamber differed from -1.4 to 2.9% for the {sup 90}Sr/{sup 90}Y and -0.3% for the {sup 85}Kr fields; their angular variation showed differences lower than 2% for incidence angles up to 40-degrees and it reached 11% for higher angles, when compared to ISO values. Measurements with the radiochromic film showed an asymmetry of the radiation field that is caused by a misalignment. Differences between the angular variations of absorbed dose rates determined by both dosimetry systems suggested that some correction factors for the extrapolation chamber that were not considered should be determined. (Author)

  13. Poster – 13: Evaluation of an in-house CCD camera film dosimetry imaging system for small field deliveries

    Energy Technology Data Exchange (ETDEWEB)

    Lalonde, Michel; Alexander, Kevin; Olding, Tim; Schreiner, L. John; Kerr, Andrew T. [Cancer Centre of Southeastern Ontario at KGH, Queen’s University (Canada)

    2016-08-15

    Purpose: Radiochromic film dosimetry is a standard technique used in clinics to verify modern conformal radiation therapy delivery, and sometimes in research to validate other dosimeters. We are using film as a standard for comparison as we improve high-resolution three-dimensional gel systems for small field dosimetry; however, precise film dosimetry can be technically challenging. We report here measurements for fractionated stereotactic radiation therapy (FSRT) delivered using volumetric modulated arc therapy (VMAT) to investigate the accuracy and reproducibility of film measurements with a novel in-house readout system. We show that radiochromic film can accurately and reproducibly validate FSRT deliveries and also benchmark our gel dosimetry work. Methods: VMAT FSRT plans for metastases alone (PTV{sub MET}) and whole brain plus metastases (WB+PTV{sub MET}) were delivered onto a multi-configurational phantom with a sheet of EBT3 Gafchromic film inserted mid-plane. A dose of 400 cGy was prescribed to 4 small PTV{sub MET} structures in the phantom, while a WB structure was prescribed a dose of 200 cGy in the WB+PTV{sub MET} iterations. Doses generated from film readout with our in-house system were compared to treatment planned doses. Each delivery was repeated multiple times to assess reproducibility. Results and Conclusions: The reproducibility of film optical density readout was excellent throughout all experiments. Doses measured from the film agreed well with plans for the WB+PTV{sub MET} delivery. But, film doses for PTV{sub MET} only deliveries were significantly below planned doses. This discrepancy is due to stray/scattered light perturbations in our system during readout. Corrections schemes will be presented.

  14. Poster – 13: Evaluation of an in-house CCD camera film dosimetry imaging system for small field deliveries

    International Nuclear Information System (INIS)

    Lalonde, Michel; Alexander, Kevin; Olding, Tim; Schreiner, L. John; Kerr, Andrew T.

    2016-01-01

    Purpose: Radiochromic film dosimetry is a standard technique used in clinics to verify modern conformal radiation therapy delivery, and sometimes in research to validate other dosimeters. We are using film as a standard for comparison as we improve high-resolution three-dimensional gel systems for small field dosimetry; however, precise film dosimetry can be technically challenging. We report here measurements for fractionated stereotactic radiation therapy (FSRT) delivered using volumetric modulated arc therapy (VMAT) to investigate the accuracy and reproducibility of film measurements with a novel in-house readout system. We show that radiochromic film can accurately and reproducibly validate FSRT deliveries and also benchmark our gel dosimetry work. Methods: VMAT FSRT plans for metastases alone (PTV MET ) and whole brain plus metastases (WB+PTV MET ) were delivered onto a multi-configurational phantom with a sheet of EBT3 Gafchromic film inserted mid-plane. A dose of 400 cGy was prescribed to 4 small PTV MET structures in the phantom, while a WB structure was prescribed a dose of 200 cGy in the WB+PTV MET iterations. Doses generated from film readout with our in-house system were compared to treatment planned doses. Each delivery was repeated multiple times to assess reproducibility. Results and Conclusions: The reproducibility of film optical density readout was excellent throughout all experiments. Doses measured from the film agreed well with plans for the WB+PTV MET delivery. But, film doses for PTV MET only deliveries were significantly below planned doses. This discrepancy is due to stray/scattered light perturbations in our system during readout. Corrections schemes will be presented.

  15. EBT-S 28-GHz, 200-kW, CW, mixed-mode, quasi-optical plasma heating system

    International Nuclear Information System (INIS)

    White, T.L.; Kimrey, H.D.; Bigelow, T.S.; Bates, D.D.; Eason, H.O.

    1984-07-01

    The ELMO Bumpy Torus-Scale (EBT-S) 28-GHz, 200-kW, cw, plasma heating system consists of a gyrotron oscillator, an oversized waveguide two-bend transmission system, and a quasi-optical mixed-mode microwave distribution manifold that feeds microwave power to the 24 plasma loads of the EBT-S fusion experiment. Balancing power to the 24 loads of the EBT-S fusion experiment. Balancing power to the 24 loads was achieved by adjusting the areas at 24 coupling irises. System performance is easily measured using system calorimetry. The distribution manifold mixed-mode power transmission, reflection, and loss coefficients are 89%, 6%, and 5%, respectively. The overall system efficiency (plasma power/gyrotron power) is 80%, but with some modifications to the distribution manifold we believe the ultimate efficiency can approach 90%. The system reliability is outstanding with a world's record 1 x 10 5 kW h of 28-GHz energy delivered to the EBT-S device with well over 1 x 10 3 operating hours

  16. Dosimetric verification of stereotactic radiosurgery/stereotactic radiotherapy dose distributions using Gafchromic EBT3

    Energy Technology Data Exchange (ETDEWEB)

    Cusumano, Davide, E-mail: davide.cusumano@unimi.it [School of Medical Physics, University of Milan, Milan (Italy); Fumagalli, Maria L. [Health Department, Fondazione IRCCS Istituto Neurologico Carlo Besta, Milan (Italy); Marchetti, Marcello; Fariselli, Laura [Department of Neurosurgery, Radiotherapy Unit, Fondazione IRCCS Istituto Neurologico Carlo Besta, Milan (Italy); De Martin, Elena [Health Department, Fondazione IRCCS Istituto Neurologico Carlo Besta, Milan (Italy)

    2015-10-01

    Aim of this study is to examine the feasibility of using the new Gafchromic EBT3 film in a high-dose stereotactic radiosurgery and radiotherapy quality assurance procedure. Owing to the reduced dimensions of the involved lesions, the feasibility of scanning plan verification films on the scanner plate area with the best uniformity rather than using a correction mask was evaluated. For this purpose, signal values dispersion and reproducibility of film scans were investigated. Uniformity was then quantified in the selected area and was found to be within 1.5% for doses up to 8 Gy. A high-dose threshold level for analyses using this procedure was established evaluating the sensitivity of the irradiated films. Sensitivity was found to be of the order of centiGray for doses up to 6.2 Gy and decreasing for higher doses. The obtained results were used to implement a procedure comparing dose distributions delivered with a CyberKnife system to planned ones. The procedure was validated through single beam irradiation on a Gafchromic film. The agreement between dose distributions was then evaluated for 13 patients (brain lesions, 5 Gy/die prescription isodose ~80%) using gamma analysis. Results obtained using Gamma test criteria of 5%/1 mm show a pass rate of 94.3%. Gamma frequency parameters calculation for EBT3 films showed to strongly depend on subtraction of unexposed film pixel values from irradiated ones. In the framework of the described dosimetric procedure, EBT3 films proved to be effective in the verification of high doses delivered to lesions with complex shapes and adjacent to organs at risk.

  17. Calibration of GafChromic EBT3 for absorbed dose measurements in 5 MeV proton beam and 60Co γ-rays

    International Nuclear Information System (INIS)

    Vadrucci, M.; Ronsivalle, C.; Marracino, F.; Montereali, R. M.; Picardi, L.; Piccinini, M.; Vincenti, M. A.; Esposito, G.; De Angelis, C.; Cherubini, R.; Pimpinella, M.

    2015-01-01

    Purpose: To study EBT3 GafChromic film in low-energy protons, and for comparison purposes, in a reference 60 Co beam in order to use it as a calibrated dosimetry system in the proton irradiation facility under construction within the framework of the Oncological Therapy with Protons (TOP)-Intensity Modulated Proton Linear Accelerator for RadioTherapy (IMPLART) Project at ENEA-Frascati, Italy. Methods: EBT3 film samples were irradiated at the Istituto Nazionale di Fisica Nucleare—Laboratori Nazionali di Legnaro, Italy, with a 5 MeV proton beam generated by a 7 MV Van de Graaff CN accelerator. The nominal dose rates used were 2.1 Gy/min and 40 Gy/min. The delivered dose was determined by measuring the particle fluence and the energy spectrum in air with silicon surface barrier detector monitors. A preliminary study of the EBT3 film beam quality dependence in low-energy protons was conducted by passively degrading the beam energy. EBT3 films were also irradiated at ENEA-National Institute of Ionizing Radiation Metrology with gamma radiation produced by a 60 Co source characterized by an absorbed dose to water rate of 0.26 Gy/min as measured by a calibrated Farmer type ionization chamber. EBT3 film calibration curves were determined by means of a set of 40 film pieces irradiated to various doses ranging from 0.5 Gy to 30 Gy absorbed dose to water. An EPSON Expression 11000XL color scanner in transmission mode was used for film analysis. Scanner response stability, intrafilm uniformity, and interfilm reproducibility were verified. Optical absorption spectra measurements were performed on unirradiated and irradiated EBT3 films to choose the most sensitive color channel to the dose range used. Results: EBT3 GafChromic films show an under response up to about 33% for low-energy protons with respect to 60 Co gamma radiation, which is consistent with the linear energy transfer dependence already observed with higher energy protons, and a negligible dose-rate dependence in

  18. Energy deposition evaluation for ultra-low energy electron beam irradiation systems using calibrated thin radiochromic film and Monte Carlo simulations

    Energy Technology Data Exchange (ETDEWEB)

    Matsui, S., E-mail: smatsui@gpi.ac.jp; Mori, Y. [The Graduate School for the Creation of New Photonics Industries, 1955-1 Kurematsucho, Nishiku, Hamamatsu, Shizuoka 431-1202 (Japan); Nonaka, T.; Hattori, T.; Kasamatsu, Y.; Haraguchi, D.; Watanabe, Y.; Uchiyama, K.; Ishikawa, M. [Hamamatsu Photonics K.K. Electron Tube Division, 314-5 Shimokanzo, Iwata, Shizuoka 438-0193 (Japan)

    2016-05-15

    For evaluation of on-site dosimetry and process design in industrial use of ultra-low energy electron beam (ULEB) processes, we evaluate the energy deposition using a thin radiochromic film and a Monte Carlo simulation. The response of film dosimeter was calibrated using a high energy electron beam with an acceleration voltage of 2 MV and alanine dosimeters with uncertainty of 11% at coverage factor 2. Using this response function, the results of absorbed dose measurements for ULEB were evaluated from 10 kGy to 100 kGy as a relative dose. The deviation between the responses of deposit energy on the films and Monte Carlo simulations was within 15%. As far as this limitation, relative dose estimation using thin film dosimeters with response function obtained by high energy electron irradiation and simulation results is effective for ULEB irradiation processes management.

  19. SU-E-T-476: Quality Assurance for Gamma Knife Perfexion Using the Exradin W1 Plastic Scintillation Detector

    Energy Technology Data Exchange (ETDEWEB)

    Pino, R [Houston Methodist Hospital (United States); Therriault-Proulx, F; Yang, J; Beddar, S [University of Texas MD Anderson Cancer Center, Houston, TX (United States)

    2014-06-01

    Purpose: To perform dose profile and output factor measurements for the Exradin W1 plastic scintillation detector (PSD) for the Gamma Knife Perfexion (GKP) collimators in a Lucy phantom and to compare these values to an Exradin A16 ion chamber, EBT3 radiochromic film and treatment planning system (TPS) data. Methods: We used the Exradin W1 PSD which has a small volume, near-water equivalent sensitive element. It has also been shown to be energy independent. This new detector is manufactured and distributed by Standard Imaging, Inc. Measurements were performed for all three collimators (4 mm, 8 mm and 16 mm) for the GKP. The Lucy phantom with the PSD inserted was moved in small steps to acquire profiles in all three directions. EBT3 film was inserted in the Lucy phantom and exposed to a single shot for each collimator. Relative output factors were measured using the three detectors while profiles acquired with the PSD were compared to the ones measured with EBT3 radiochromic film. Results: Measured output factors relative to the largest collimator are as followsCollimator PS EBT3 A1616mm 1.000 1.000 1.0008mm 0.892 0.881 0.8834mm 0.795 0.793 0.727 The nominal (vendor) OFs for GKP are 1.000, 0.900, and 0.814, for collimators 16 mm, 8 mm and 4 mm, respectively. There is excellent agreement between all profiles measured with the PSD and EBT3 as well as with the TPS data provided by the vendor. Conclusion: Output factors measured with the W1 were consistent with the ones measured with EBT3 and A16 ion chamber. Measured profiles are in excellent agreement. The W1 detector seems well suited for beam QA for Gamma Knife due to its dosimetric characteristics. Sam Beddar would like to disclose a NIH/NCI SBIR Phase II grant (2R44CA153824-02A1) with Standard Imaging, Title: “Water-Equivalent Plastic Scintillation Detectors for Small Field Radiotherapy”.

  20. Absolute and relative dose measurements with Gafchromic trade mark sign EBT film for high energy electron beams with different doses per pulse

    International Nuclear Information System (INIS)

    Fiandra, Christian; Ragona, Riccardo; Ricardi, Umberto; Anglesio, Silvia; Giglioli, Francesca Romana

    2008-01-01

    The authors have evaluated the accuracy, in absolute and relative dose measurements, of the Gafchromic trade mark sign EBT film in pulsed high-energy electron beams. Typically, the electron beams used in radiotherapy have a dose-per-pulse value of less than 0.1 mGy/pulse. However, very high dose-per-pulse electron beams are employed in certain linear accelerators dedicated to intraoperatory radiation therapy (IORT). In this study, the absorbed dose measurements with Gafchromic trade mark sign EBT in both low (less than 0.3 mGy per pulse) and high (30 and 70 mGy per pulse) dose-per-pulse electron beams were compared with ferrous sulfate chemical Fricke dosimetry (operated by the Italian Primary Standard Dosimetry Laboratory), a method independent of the dose per pulse. A summary of Gafchromic trade mark sign EBT in relative and absolute beam output determination is reported. This study demonstrates the independence of Gafchromic trade mark sign EBT absorption as a function of dose per pulse at different dose levels. A good agreement (within 3%) was found with Fricke dosimeters for plane-base IORT applicators. Comparison with a diode detector is presented for relative dose measurements, showing acceptable agreement both in the steep dose falloff zone and in the homogeneous dose region. This work also provides experimental values for recombination correction factor (K sat ) of a Roos (plane parallel) ionization chamber calculated on the basis of theoretical models for charge recombination.

  1. Isodose curves recorded in radiochromic film of an iodine seed array

    International Nuclear Information System (INIS)

    Mourao, Arnaldo Prata; Tomaz, Lucas Crusoe; Grynberg, Suely Epzstein

    2015-01-01

    Brachytherapy seeds are frequently used in cancer treatment. Iodine seeds are ranked among the top choices when it comes to the treatment of prostate cancer. Iodine-125 emits X and gamma photons with an average energy of 28 keV and a half-life of 59.4 days. A set of four iodine-125 seeds, model 6711 produced by Amersham Health, were used in this work. The dosimetric characteristics for a seed were obtained taking into account the recommendations of TG-43 protocol, developed by the AAPM (American Association of Physicists in Medicine). To realize the experiment three plates of Standard Grade Solid Water, model 457 ® Gammex were used. One solid water plate was machined for accommodate the seed set. The set of seeds was placed in a symmetrical configuration trying to simulate an arrangement which may occur in vivo during treatment. A radiochromic film sheet was placed on the plate machined after the seed charge. The machined plate charged and with the film sheet was placed between two others solid water plate to recorder the radiation dose. The machined plate was placed between the other two plates for better reliability in measurements. The radio chromic film was irradiated by an equivalent X-ray beam using the reference radiation RQR 3 IEC (International Electrotechnical Commission) with different doses (0.5 to 1.0 Gy) to obtain the calibration curve in the dose region measured. After validation of the methodology, the study of the interaction between the radiation fields of the set of seeds became possible and the survey of isodose curves of these setting was conducted. (author)

  2. EBT-S ECH transmission systems

    International Nuclear Information System (INIS)

    White, T.

    1983-01-01

    Over the past 18 months the EBT-S 28-GHz 200 kW cw mixed mode ECH system efficiency has increased from 70 to 80% by the substitution of 90 0 smooth radius W/G bends for the earlier miter bends, by eliminating several mode absorbers necessary for miter bend operation, and by optimizing the microwave coupling through the quasi-optical toroidal distribution manifold. Along with reduced system losses, these improvements have allowed the gyrotron sources to operate in a wider parameter range with high dc to rf conversion efficiency than previously attainable. Other components that have been developed besides radiused bends include T/sub i/O 2 mode absorbers, power splitters, bellows, and several types of high field mixed mode launchers. Details on these components will be presented as time allows

  3. Electron confinement studies on the EBT-S Bumpy Torus Experiment using soft x-ray techniques

    International Nuclear Information System (INIS)

    Hillis, D.L.; Haste, G.R.; Berry, L.A.

    1983-01-01

    Soft x-ray bremsstrahlung measurements have been performed on the ELMO Bumpy Torus (EBT-S) plasma to determine the electron temperature T/sub e/ and electron density density n/sub e/ using a calibrated Si(Li) detector over a wide range of operating conditions. The purpose of this paper is to outline the necessary assumptions and essential x-ray techniques that are inherent in soft x-ray measurements in order to investigate the electron heating and confinement properties of EBT-S. In addition, by utilizing the electron density as determined by the soft x-ray measurements, the previous EBT-S confinement analyses have been extended. The steady-state plasma of EBT-S is heated by microwaves using a continuous wave (cw) gyrotron that can operate up to power levels of 200 kW. From the soft x-ray measurements, both the electron temperature and density are found to increase at higher microwave power levels. For operation at microwave power levels of 200 kW, T/sub e/ approaches 1 keV while n/sub e/ approaches 1.2 x 10 12 cm - 3 . In general, confinement properties are found to improve with increased microwave power. The data are compared with neoclassical transport scaling and the electron transport is found to be collisionless (#betta#/Ω<1) as well as neoclassical

  4. A standard comparison of spectral properties and energy response for a number of ultraviolet dosimeters

    International Nuclear Information System (INIS)

    Abu Kassem, I.; Bero, M.

    2013-09-01

    The radio-chromic chemical radiation dosimeters are materials that change their optical properties (Spectrique optical absorbance and optical density) due to radiation absorption. These detectors are used for dose measurements of ionizing radiation (γ and X-rays). It was established that these detectors could be used for ultraviolet radiation dose measurements. So, the aim of this work is to study and compare the metrological properties of two radiation chemical detectors, FXG gel and EBT2 Gafchromic film. The FXG gel is prepared directly at laboratory but EBT2 gafchromic film is a self developed radiation sensible film which is available as commercial product. This work focuses on realizing a comparison study between FXG and EBT2 detector metrological properties. It consists of studying optical and spectral properties of the detectors responsivity, radiation and temporal stability, linearity and total detected dose level. The results showed that the studied detectors present a very good responsivity to UVA radiation, high stability in optical absorbance under UVA radiation and good linearity over wide radiation level which contains the solar UVA radiation level reaching the earth surface. But, the EBT2 film presents two time higher total detection dose level than FXG gel, moreover, due to the simplicity of use, it was possible to test the use of EBT2 film for direct solar UVA radiation measurement. The two studied chemical detector (FXG gel and EBT2 film) provide a direct second order mathematical relation between the applied radiation dose and the optical absorbance changes with a very good approximation and suitable uncertainty (Measurement results relative dispersion is about 5%). It is possible to study the EBT2 film optical density variation as a function of UVA dose using directly a portable densitometer (author).

  5. Development and clinical implementation of a global software use and analysis of radiochromic s films; Desarrollo e implementacion clinica de un software global de uso y analisis de peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Gimeno Olmos, J.; Martinez Fernandez, J. M.; Pastor Sanchis, V.; Gonzalez Perez, V.; Guardino de la Flor, C.; Dolores Alemany, V. de los; Crispin Contreras, V.

    2011-07-01

    The aim of this study was to gather information in one solution the different processes that occur with the use of radiochromic films: calibration, reading, corrections, step-absorbed dose, dose distribution analysis and comparison with reference distributions.

  6. SU-E-T-71: A Radiochromic Film Based Quantitative Assessment of Thermoplastic Mask Bolus Effect in Head and Neck IMRT/VMAT

    Energy Technology Data Exchange (ETDEWEB)

    Kalavagunta, C; Lin, M; Snider, J; Xu, H; Schrum, A; Vadnais, P; Marter, K; Suntharalingam, M; Prado, K [University of Maryland School of Medicine, Baltimore, MD (United States)

    2015-06-15

    Purpose: To quantify the factors leading to thermoplastic mask bolus-associated-increased skin dose in head and neck IMRT/VMAT using EBT2 film. Methods: EBT2 film placed beneath a dual layer 3-point ORFIT head, neck and shoulder mask was used to test the effect of mask thickness, beam modulation, air gap, and beam obliquity on bolus effect. Mask thickness was varied based on the distribution of 1.6mm Orfilight layer on top of 2 mm Efficast layer. Beam modulation was varied by irradiating the film with an open field (no beam modulation) and a step and shoot field (beam modulation). Air gap between mask and film was varied from 0 to 5mm. Beam obliquity was varied by irradiating the film at gantry angles of 0°, 35°, and 70°.Finally, film strips placed on a Rando phantom under an Orfit mask, in regions of expected high dose, were irradiated using 5 IMRT and 5 VMAT plans with various modulation levels (modulation factor 2 to 5) and the results were compared with those obtained placing OSLDs at the same locations. Results: An 18–34% increase in mask bolus effect was observed for three factors where the effect of beam obliquity ≥ beam modulation > mask thickness. No increase in mask bolus effect was observed for change in air gap. A 6–13% increase in dose due to mask bolus effect was observed on film strips. Conclusion: This work underlines the role of beam obliquity and beam modulation combined with thermoplastic mask thickness in increasing mask bolus-associated skin dose in head and neck IMRT/VMAT. One possible method of dose reduction, based on knowledge gained from this work, is inclusion of skin as an avoidance structure in treatment planning. Another approach is to design a mask with the least amount of thermoplastic material necessary for immobilization.

  7. SU-E-T-71: A Radiochromic Film Based Quantitative Assessment of Thermoplastic Mask Bolus Effect in Head and Neck IMRT/VMAT

    International Nuclear Information System (INIS)

    Kalavagunta, C; Lin, M; Snider, J; Xu, H; Schrum, A; Vadnais, P; Marter, K; Suntharalingam, M; Prado, K

    2015-01-01

    Purpose: To quantify the factors leading to thermoplastic mask bolus-associated-increased skin dose in head and neck IMRT/VMAT using EBT2 film. Methods: EBT2 film placed beneath a dual layer 3-point ORFIT head, neck and shoulder mask was used to test the effect of mask thickness, beam modulation, air gap, and beam obliquity on bolus effect. Mask thickness was varied based on the distribution of 1.6mm Orfilight layer on top of 2 mm Efficast layer. Beam modulation was varied by irradiating the film with an open field (no beam modulation) and a step and shoot field (beam modulation). Air gap between mask and film was varied from 0 to 5mm. Beam obliquity was varied by irradiating the film at gantry angles of 0°, 35°, and 70°.Finally, film strips placed on a Rando phantom under an Orfit mask, in regions of expected high dose, were irradiated using 5 IMRT and 5 VMAT plans with various modulation levels (modulation factor 2 to 5) and the results were compared with those obtained placing OSLDs at the same locations. Results: An 18–34% increase in mask bolus effect was observed for three factors where the effect of beam obliquity ≥ beam modulation > mask thickness. No increase in mask bolus effect was observed for change in air gap. A 6–13% increase in dose due to mask bolus effect was observed on film strips. Conclusion: This work underlines the role of beam obliquity and beam modulation combined with thermoplastic mask thickness in increasing mask bolus-associated skin dose in head and neck IMRT/VMAT. One possible method of dose reduction, based on knowledge gained from this work, is inclusion of skin as an avoidance structure in treatment planning. Another approach is to design a mask with the least amount of thermoplastic material necessary for immobilization

  8. Radiochromic film use to record dose profile variations in chest CT scan; Utilizacao do filme radiocromico para registro da variacao do perfil de dose em varredura de TC de torax

    Energy Technology Data Exchange (ETDEWEB)

    Mourao, Arnaldo P.; Santana, Priscila C., E-mail: aprata@des.cefetmg.br [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear; Silva, Teogenes A. da; Alonso, Thessa C. [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2015-12-15

    The CT scans generate higher doses in patients than those caused by other types of diagnostic radiology using the attenuation of X-ray beams. Technological advances have increased the clinical applications of computed tomography (CT) and consequently the demand for these exams. Dose reduction strategies are difficult to implement because of a lack of proper guidance on the CT scanning. However, CT offers the possibility of adjusting the image acquisition parameters according to the patient physical profile and the diagnostic application for which it is intended to scan. Knowledge of the dose distribution is important when thinking about varying the acquisition parameters to reduce the dose. In this work was used a PMMA chest phantom to observe the dose deposition in 5 areas. Radiochromic film strips were placed inside the chest phantom, in peripheral and in the central region. The phantom was placed in the scanner isocenter and it was performed the irradiation of its central region using a chest acquisition protocol. After the phantom CT scan the radiochromic film strips were processed for obtaining digital images. Digital images were reworked to obtain the dose variation profiles for each position. The results showed a wide variation in absorbed dose by the phantom, either within a same position along the length of the film strip, as in the comparison among the five regions which the strips are placed. In this second case the dose variations were even greater. (author)

  9. Gas injection in EBT-S for assessment of particle loading effects of neutral beam injection

    International Nuclear Information System (INIS)

    Carpenter, K.H.; Glowienka, J.C.

    1979-01-01

    Experiments have begun to examine the physics of neutral beam injection on EBT-S. Preliminary experiments have been limited to a calibrated gas puffing experiment which simulates the effects of a pulsed beam with zero energy. These experiments begin to address some of the compatibility problems that exist for future beam heating experiments on EBT devices. In particular, neutral beams are to be a significant part of the planned EBT-II experiment which is designed to demonstrate steady-state, reactor-like conditions with both electron cyclotron heating and neutral beam heating

  10. Temperature, humidity and time., Combined effects on radiochromic film dosimeters

    International Nuclear Information System (INIS)

    Abdel-Fattah, A.A.; Miller, A.

    1996-01-01

    The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Riso B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60 o C for irradiation by 60 Co photons and 10-MeV electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is + 0.25 ± 0.1% per o C for the FWT-60-00 dosimeters and +0.5 ± 0.1% per o C for Riso B3 dosimeters at temperatures between 20 and 50 o C and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger dependences. Whenever possible one should use dosimeters sealed in pouches under controlled intermediate humidity conditions (30-50%) or, if that is impractical, one should maintain conditions of calibration as close as possible to the conditions of use. Without that precaution, severe dosimetry errors may result. (author)

  11. SU-F-18C-10: Clinical Implementation of Radiochromic Film Based CTDI Measurements

    International Nuclear Information System (INIS)

    Quintero, C; Seuntjens, J; Devic, S; Tomic, N; DeBlois, F

    2014-01-01

    Purpose: To evaluate accuracy and reproducibility of a radiochromic film-based protocol to measure computer tomography dose index (CTDI) as a part of annual QA on CT scanners and kV CBCT systems attached to linear accelerators. Methods: Energy dependence of Gafchromic XR-QA2(R) film model was tested over imaging beam qualities (50 – 140 kVp). Film pieces were irradiated in air to known values of air kerma in air (up to 10 cGy). Change in film reflectance was determined with an in-house written code using images produced by a flatbed document scanner. Calibration curves for each beam quality were created, and film responses for same air-kerma values were compared.Sets of film strips were placed into holes of a CTDI phantom and irradiated for several clinical scanning protocols on CT-simulators and CBCT systems. Film reflectance change was converted into dose to water and used to calculate CTDIvol values. Measured CTDIvol values were compared to tabulated CTDIvol values. Results: Average variations of ±5.2% in the mean film reflectance change were observed in the energy range of 80 to 140 keV, and 11.1% between 50 and 140 keV. The averaged measured CTDI values presented a mean variation for the same machine and protocol of 2.6%. However, measured CTDI values were in average 10% lower than tabulated CTDI values for CT-simulators, and 44% higher for CBCT systems. Conclusion: We found that in relatively broad range of beam qualities used in diagnostic radiology variation of film response is within ±5% resulting in ±15% systematic error in dose estimates if a single calibration curve is used. Relatively large discrepancy between measured and tabulated CTDI values for different protocols and imaging systems used within radiotherapy department strongly support the trend towards replacing CTDI value with equilibrium dose measurement in the center of cylindrical phantom as suggested by TG-111. This work was supported by the Natural Sciences and Engineering Research

  12. Relative dosimetry by Ebt-3; Dosimetria relativa por EBT3

    Energy Technology Data Exchange (ETDEWEB)

    De Leon A, M. A.; Rivera M, T. [IPN, Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Av. Legaria 694, 11500 Mexico D. F. (Mexico); Hernandez O, J. O., E-mail: madla16@hotmail.com [Hospital General de Mexico, Dr. Balmis 148, Col. Doctores, 06726 Mexico D. F. (Mexico)

    2015-10-15

    In the present work relative dosimetry in two linear accelerator for radiation therapy was studied. Both Varian Oncology systems named Varian Clinac 2100-Cd and MLC Varian Clinac i X were used. Gaf Chromic Ebt-3 film was used. Measurements have been performed in a water equivalent phantom, using 6 MV and 18 MV photon beams on both Linacs. Both calibration and Electron irradiations were carried out with the ionization chamber placed at the isocenter, below a stack of solid water slabs, at the depth of dose maximum (D max), with a Source-to-Surface Distance (SSD) of 100 cm and a field size of 10 cm x 10 cm. Calibration and dosimetric measurements photons were carried out under IAEA-TRS 398 protocol. Results of relative dosimetry in the present work are discussed. (Author)

  13. Development of a novel quality assurance system based on rolled-up and rolled-out radiochromic films in volumetric modulated arc therapy

    International Nuclear Information System (INIS)

    Park, Ji-Yeon; Lee, Jeong-Woo; Choi, Kyoung-Sik; Lee, Jung Seok; Kim, You-Hyun; Hong, Semie; Suh, Tae-Suk

    2011-01-01

    Purpose: To develop a cylindrical phantom with rolled-up radiochromic films and dose analysis software in the rolled-out plane for quality assurance (QA) in volumetric modulated arc therapy (VMAT). Methods: The phantom consists of an acrylic cylindrical body wrapped with radiochromic film inserted into an outer cylindrical shell of 5 cm thickness. The rolled-up films with high spatial resolution enable detection of specific dose errors along the arc trajectory of continuously irradiated and modulated beams in VMAT. The developed dose analysis software facilitates dosimetric evaluation in the rolled-up and rolled-out planes of the film; the calculated doses on the corresponding points where the rolled-up film was placed were reconstructed into a rectangular dose matrix equivalent to that of the rolled-out plane of the film. The VMAT QA system was implemented in 3 clinical cases of prostate, nasopharynx, and pelvic metastasis. Each calculated dose on the rolled-out plane was compared with measurement values by modified gamma evaluation. Detected positions of dose disagreement on the rolled-out plane were also distinguished in cylindrical coordinates. The frequency of error occurrence and error distribution were summarized in a histogram and in an axial view of rolled-up plane to intuitively identify the corresponding positions of detected errors according to the gantry angle. Results: The dose matrix reconstructed from the developed VMAT QA system was used to verify the measured dose distribution along the arc trajectory. Dose discrepancies were detected on the rolled-out plane and visualized on the calculated dose matrix in cylindrical coordinates. The error histogram obtained by gamma evaluation enabled identification of the specific error frequency at each gantry angular position. The total dose error occurring on the cylindrical surface was in the range of 5%-8% for the 3 cases. Conclusions: The developed system provides a practical and reliable QA method to

  14. Development of a novel quality assurance system based on rolled-up and rolled-out radiochromic films in volumetric modulated arc therapy

    Energy Technology Data Exchange (ETDEWEB)

    Park, Ji-Yeon; Lee, Jeong-Woo; Choi, Kyoung-Sik; Lee, Jung Seok; Kim, You-Hyun; Hong, Semie; Suh, Tae-Suk [Department of Biomedical Engineering and Research Institute of Biomedical Engineering, College of Medicine, Catholic University of Korea, Seoul 137-701 (Korea, Republic of); Research Institute of Health Science, College of Health Science, Korea University, Seoul 136-703 (Korea, Republic of) and Department of Radiation Oncology, Konkuk University Medical Center, Seoul 143-729 (Korea, Republic of); Department of Radiation Oncology, Anyang SAM Hospital, Anyang 430-733 (Korea, Republic of) and Department of Biomedical Engineering and Research Institute of Biomedical Engineering, College of Medicine, Catholic University of Korea, Seoul 137-701 (Korea, Republic of); Department of Radiation Oncology, Anyang SAM Hospital, Anyang 430-733 (Korea, Republic of); Department of Radiologic Science, College of Health Science, Korea University, Seoul 136-703 (Korea, Republic of); Department of Radiation Oncology, Konkuk University Medical Center, Seoul 143-729 (Korea, Republic of); Department of Biomedical Engineering and Research Institute of Biomedical Engineering, College of Medicine, Catholic University of Korea, Seoul 137-701 (Korea, Republic of)

    2011-12-15

    Purpose: To develop a cylindrical phantom with rolled-up radiochromic films and dose analysis software in the rolled-out plane for quality assurance (QA) in volumetric modulated arc therapy (VMAT). Methods: The phantom consists of an acrylic cylindrical body wrapped with radiochromic film inserted into an outer cylindrical shell of 5 cm thickness. The rolled-up films with high spatial resolution enable detection of specific dose errors along the arc trajectory of continuously irradiated and modulated beams in VMAT. The developed dose analysis software facilitates dosimetric evaluation in the rolled-up and rolled-out planes of the film; the calculated doses on the corresponding points where the rolled-up film was placed were reconstructed into a rectangular dose matrix equivalent to that of the rolled-out plane of the film. The VMAT QA system was implemented in 3 clinical cases of prostate, nasopharynx, and pelvic metastasis. Each calculated dose on the rolled-out plane was compared with measurement values by modified gamma evaluation. Detected positions of dose disagreement on the rolled-out plane were also distinguished in cylindrical coordinates. The frequency of error occurrence and error distribution were summarized in a histogram and in an axial view of rolled-up plane to intuitively identify the corresponding positions of detected errors according to the gantry angle. Results: The dose matrix reconstructed from the developed VMAT QA system was used to verify the measured dose distribution along the arc trajectory. Dose discrepancies were detected on the rolled-out plane and visualized on the calculated dose matrix in cylindrical coordinates. The error histogram obtained by gamma evaluation enabled identification of the specific error frequency at each gantry angular position. The total dose error occurring on the cylindrical surface was in the range of 5%-8% for the 3 cases. Conclusions: The developed system provides a practical and reliable QA method to

  15. Dual-peak dose measurement for radiochromic films by a newly developed spectral microdensitometer

    International Nuclear Information System (INIS)

    Lee, K.Y.; Fung, K.K.L.; Kwok, C.S.

    2005-01-01

    Radiochromic film (RCF) dosimetry is usually based on densitometric methods which use an analyzing light source of a fixed or a broad spectrum of wavelengths. These methods have not exploited the sensitivity of the dose response of the RCF otherwise attainable by using a light source with wavelengths peaked at the two absorption peaks in the absorption spectrum of the RCF. A new algorithm of dual-peak dose measurement for the RCF has been proposed in this paper to make use of these dual absorption peaks to achieve the maximum attainable sensitivity. This technique relies on the measurement of the transmittance of the RCF at the wavelength of the major and minor absorption peaks, respectively. The dual-peak dose measurement is accomplished with the aid of a novel spectral microdensitometer developed in our Institute. The microdensitometer utilizes a monochromator to provide a light source of which the wavelength can be matched precisely to the wavelength of the absorption peaks of the RCF. The doses obtained at these wavelengths are fed into a weighted objective function and an optimum dose is searched by minimizing the objective function to give the best estimate of the dose deposited on the film. An initial test shows that there is a good agreement between the estimated and actual dose deposited; and the maximum discrepancy was found to be less than 1%

  16. Characterization of a team intraoperative Radiation therapy and measurement of dose in skin with film radiochromic

    International Nuclear Information System (INIS)

    Onses Segarra, A.; Sancho Kolster, I.; Eraso Urien, A.; Pla Farnos, M. J.; Picon Olmos, C.

    2015-01-01

    This paper presents the results of the initial reference state of intraoperative radiotherapy equipment lntraBeam, for performing breast treatments are analyzed. To the initial reference team was established for the following dosimetric and geometric beam parameters: percentage depth dose, beam quality, isotropy, linearity and mechanical and geometric integrity for both the source RX as for different spherical applicators of the team. Based on these checks, a program of periodic quality control was established. One of the exclusion criteria for this treatment is that the tumor is less than l cm of the skin, yaque give doses received in this organ can be high. For this reason it is important to know exactly the absorbed dose in skin during these treatments. In this regard we have implemented a system for measuring the skin dose during treatment with Radiochromic film, placing 4 film segments in fixed positions of the skin around the surgical incision. It .ha obtained calibration curve of sterilized films and compared the results with a calibration beam megavoltage. The results of the skin dose measurements are compared with theoretical estimates given by the planning system equipment. The results indicate the need to measure individually the skin dose for these treatments. (Author)

  17. New radiochroic film densitometry system

    International Nuclear Information System (INIS)

    Tran, T.; Freeman, N.; Johnston, P.

    2004-01-01

    Full text: The advantages of radiochromic film in radiation dosimetry are well known. They include dosimetry with high spatial resolution, response less dependent on incident beam energy than common radiotherapy films (such as the Kodak XV films), tissue equivalence and the ability to be handled and developed in room light. This study entails the design and testing of a new radiochromic densitometry system. The system consists of a single light emitting diode (LED), opaque 'diffuser' and digital camera. Customised software was developed to analyse images obtained from the digital camera. Standard characteristics of a commercially available super bright red LED (peak wavelength 625nm) was analysed in order to determine the voltage, current and intensity settings. Various methods in diffusing the single LED light source were investigated and it was determined that an opaque transmission 'diffuser' was the best alternative. While the intensity of the LED was kept constant, the digital camera exposure times were varied in order to determine a setting which would produce the best image exposure. The system was designed and built and preliminary tests were carried using the standard radiochromic film GafChromic MD-55-2. LED current vs. voltage curves were characteristically exponential for positive voltage. Studies into LED intensity versus camera exposure produced an unexpected result. At high exposures the camera saturates and if even higher exposures are used the LED intensity apparently decreases. This was thought to be due to the pixels in the charge couple device (CCD) saturating and eventually electronically 'bleeding' into adjacent pixels. Using the opaque transmission 'diffuser' enabled successful use of the single LED light producing an area of homogenous light intensity in which images of radiochromic films can be obtained. Preliminary results from radiochromic film characteristic studies show no unusual results. The single LED, diffuser and digital camera

  18. Cyberknife Relative Output Factor measurements using fiber-coupled luminescence, MOSFETS and RADPOS dosimetry system

    DEFF Research Database (Denmark)

    Ploquin, N.; Kertzscher Schwencke, Gustavo Adolfo Vladimir; Vandervoort, E.

    2012-01-01

    from 5 to 60 mm. ROFs were also measured using a mobileMOSFET system (Best Medical Canada) and EBT1 and EBT2 GAFCHROMIC® (ISP, Ashland) radiochromic films. For cone sizes 12.5–60 mm all detector results were in agreement within the measurement uncertainty. The microMOSFET/RADPOS measurements (published.......3% and 0.865 ± 0.3% for 5, 7.5 and 10 mm cones. Our study shows that the microMOSFET/RADPOS and optical fiber‐coupled RL dosimetry system are well suited for Cyberknife cone output factors measurements over the entire range of field sizes, provided that appropriate correction factors are applied...

  19. Dose measurement using radiochromic lms and Monte Carlo simulation for hadron-therapy

    International Nuclear Information System (INIS)

    Zahra, N.

    2010-06-01

    Because of the increase in dose at the end of the range of ions, dose delivery during patient treatment with hadron-therapy should be controlled with high precision. Monte Carlo codes are now considered mandatory for validation of clinical treatment planning and as a new tool for dosimetry of ion beams. In this work, we aimed to calculate the absorbed dose using Monte Carlo simulation Geant4/Gate. The effect on the dose calculation accuracy of different Geant4 parameters has been studied for mono-energetic carbon ion beams of 300 MeV/u in water. The parameters are: the production threshold of secondary particles and the maximum step limiter of the particle track. Tolerated criterion were chosen to meet the precision required in radiotherapy in term of value and dose localisation (2%, 2 mm respectively) and to obtain the best compromise on dose distribution and computational time. We propose here the values of parameters in order to satisfy the precision required. In the second part of this work, we study the response of radiochromic films MD-v2-55 for quality control in proton and carbon ion beams. We have particularly observed and studied the quenching effect of dosimetric films for high LET (≥20 keV/μm) irradiation in homogeneous and heterogeneous media. This effect is due to the high ionization density around the track of the particle. We have developed a method to predict the response of radiochromic films taking into account the saturation effect. This model is called the RADIS model for 'Radiochromic films Dosimetry for Ions using Simulations'. It is based on the response of films under photon irradiations and the saturation of films due to high linear energy deposit calculated by Monte Carlo. Different beams were used in this study and aimed to validate the model for hadron-therapy applications: carbon ions, protons and photons at different energies. Experiments were performed at Grand Accelerateur National d'Ions Lourds (GANIL), Proton therapy center of

  20. Improved spectral data unfolding for radiochromic film imaging spectroscopy of laser-accelerated proton beams

    Energy Technology Data Exchange (ETDEWEB)

    Schollmeier, M.; Geissel, M.; Sefkow, A. B. [Sandia National Laboratories, Albuquerque, New Mexico 87185 (United States); Flippo, K. A. [Los Alamos National Laboratory, Los Alamos, New Mexico 87545 (United States)

    2014-04-15

    An improved method to unfold the space-resolved proton energy distribution function of laser-accelerated proton beams using a layered, radiochromic film (RCF) detector stack has been developed. The method takes into account the reduced RCF response near the Bragg peak due to a high linear energy transfer (LET). This LET dependence of the active RCF layer has been measured, and published data have been re-interpreted to find a nonlinear saturation scaling of the RCF response with stopping power. Accounting for the LET effect increased the integrated particle yield by 25% after data unfolding. An iterative, analytical, space-resolved deconvolution of the RCF response functions from the measured dose was developed that does not rely on fitting. After the particle number unfold, three-dimensional interpolation is performed to determine the spatial proton beam distribution for proton energies in-between the RCF data points. Here, image morphing has been implemented as a novel interpolation method that takes into account the energy-dependent, changing beam topology.

  1. Quality assurance for an image-guided frameless radiosurgery system using radiochromic film

    International Nuclear Information System (INIS)

    Shiomi, Hiroya; Inoue, Takehiro; Nakamura, Satoaki; Inoue, Toshihiko

    2000-01-01

    The CyberKnife, a new type of machine for stereotactic irradiation, is composed of a combination of a robot (manipulator) and a linear accelerator. We measured the total pointing error of the therapy beam, aiming at improvement in the accuracy of the CyberKnife. The CyberKnife keeps a number of coordinate systems such as the patient's coordinates, robotic coordinates, and IPS (image processing system) coordinates. Precise irradiation is possible if these coordinate systems are matched accurately. We always calculate the overall irradiation error using GAF MD-55 radiochromic film before irradiation treatment of the patient, and we have attempted to improve the accuracy of irradiation with the CyberKnife by feeding back the errors. The median of the error immediately after introducing the CyberKnife was 1.2 mm, and as a result of correction, we succeeded in reducing the error to 0.7 mm (median). Total pointing error after correction was significantly lower than that before correction (p=0.0023). This approach allowed us to reduce errors and establish a method for providing patients with highly accurate stereotactic irradiation. We believe these results endorse the validity of the method presented in this paper. (author)

  2. Comparative Analysis of Different Measurement Techniques for MLC Characterization: Preliminary Results

    International Nuclear Information System (INIS)

    Larraga-Gutierrez, J. M.; Ballesteros-Zebadua, P.; Garcia-Garduno, O. A.; Martinez-Davalos, A.; Rodriguez-Villafuerte, M.; Moreno-Jimenez, S.; Celis, M. A.

    2008-01-01

    Radiation transmission, leakage and beam penumbra are essential dosimetric parameters related to the commissioning of a multileaf collimation system. This work shows a comparative analysis of commonly used film detectors: X-OMAT V2 and EDR2 radiographic films, and GafChromic EBT registered radiochromic film. The results show that X-OMAT over-estimates radiation leakage and 80-20% beam penumbra. However, according to the reference values reported by the manufacturer for these dosimetric parameters, all three films are adequate for MLC dosimetric characterization, but special care must be taken when X-OMAT V2 film is used due to its low energy photon dependence

  3. Experimental dosimetry in conformal breast teletherapy compared with the planning system

    International Nuclear Information System (INIS)

    Nogueira, Luciana Batista; Silva, Hugo Leonardo Lemos; Passos Ribeiro de Campos, Tarcísio

    2015-01-01

    The objective of this study was to compare and analyse the absorbed dose profiles from the conformal radiotherapy planning and experimental dosimetry taken in a breast anthropomorphic and anthropometric phantom. Conformal radiotherapy planning was elaborated in the Treatment Planning System (TPS). EBT2 Gafchromic radiochromic films were applied as dosimeters, positioned internally and superficially in the breast phantom. The standard radiation protocol was applied in the breast phantom. The films were digitalised, and their responses were analysed in RGB. The optical densities were processed, reproducing the spatial dose distribution. - Highlights: • Distributions of absorbed doses were generated by the TPS and measured by radiochromic films. • The breast phantom simulated a human breast in position for treatment. • A large portion of the glandular tissue absorbed doses that were equivalent to the radiotherapy planning. • There were regions of hot spots and small areas of under dosage in deeper areas at the lung interface

  4. Proposed magnet system for EBT-P

    International Nuclear Information System (INIS)

    Ballou, J.K.; McManany, T.J.; Power, W.H.; Yeh, H.T.

    1979-01-01

    Since EBT-P, and other EBT devices, depends on having closed field lines for its successful operation, a very tight tolerance (on the order of 6 parts in 10 4 ) must be imposed on certain components of the field error. This is a global tolerance that influences the design of each magnet, its leads, the electrical bus, and the relative position of each magnet to its neighbors. This paper discusses the design, and field error analysis of the ORNL reference design for EBT-P

  5. Development of computational models for the simulation of isodose curves on dosimetry films generated by iodine-125 brachytherapy seeds

    International Nuclear Information System (INIS)

    Santos, Adriano M.; Meira-Belo, Luiz C.; Reis, Sergio C.; Grynberg, Suely E.

    2011-01-01

    The interstitial brachytherapy is one modality of radiotherapy in which radioactive sources are placed directly in the region to be treated or close to it. The seeds that are used in the treatment of prostate cancer are generally cylindrical radioactive sources, consisting of a ceramic or metal matrix, which acts as the carrier of the radionuclide and as the X-ray marker, encapsulated in a sealed titanium tube. This study aimed to develop a computational model to reproduce the film-seed geometry, in order to obtain the spatial regions of the isodose curves produced by the seed when it is put over the film surface. The seed modeled in this work was the OncoSeed 6711, a sealed source of iodine-125, which its isodose curves were obtained experimentally in previous work with the use of dosimetric films. For the films modeling, compositions and densities of the two types of dosimetric films were used: Agfa Personal Monitoring photographic film 2/10, manufactured by Agfa-Geavaert; and the model EBT radiochromic film, by International Specialty Products. The film-seed models were coupled to the Monte Carlo code MCNP5. The results obtained by simulations showed to be in good agreement with experimental results performed in a previous work. This indicates that the computational model can be used in future studies for other seeds models. (author)

  6. Development of computational models for the simulation of isodose curves on dosimetry films generated by iodine-125 brachytherapy seeds

    Energy Technology Data Exchange (ETDEWEB)

    Santos, Adriano M.; Meira-Belo, Luiz C.; Reis, Sergio C.; Grynberg, Suely E., E-mail: amsantos@cdtn.b [Center for Development of Nuclear Technology (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2011-07-01

    The interstitial brachytherapy is one modality of radiotherapy in which radioactive sources are placed directly in the region to be treated or close to it. The seeds that are used in the treatment of prostate cancer are generally cylindrical radioactive sources, consisting of a ceramic or metal matrix, which acts as the carrier of the radionuclide and as the X-ray marker, encapsulated in a sealed titanium tube. This study aimed to develop a computational model to reproduce the film-seed geometry, in order to obtain the spatial regions of the isodose curves produced by the seed when it is put over the film surface. The seed modeled in this work was the OncoSeed 6711, a sealed source of iodine-125, which its isodose curves were obtained experimentally in previous work with the use of dosimetric films. For the films modeling, compositions and densities of the two types of dosimetric films were used: Agfa Personal Monitoring photographic film 2/10, manufactured by Agfa-Geavaert; and the model EBT radiochromic film, by International Specialty Products. The film-seed models were coupled to the Monte Carlo code MCNP5. The results obtained by simulations showed to be in good agreement with experimental results performed in a previous work. This indicates that the computational model can be used in future studies for other seeds models. (author)

  7. Two dimensional Raman mapping with respect to carbon bonds of radiochromic films: An approach to micro-dosimetry

    International Nuclear Information System (INIS)

    Heo, Taemin; Park, Hyeonsuk; Ye, Sung-Joon

    2015-01-01

    Raman spectroscopy usually provides fingerprints of chemical component species and molecular motion. Raman peak intensity can be quantified as dose changes. Using that Raman peak intensity is proportional to the electric field intensity of incidence beam and the concentration of compounds, the dose trend would have the linearity with the concentration change of radio-active compounds. Raman spectroscopy has been applied to be utilized as a dosimetry in our group in the previous study. Then, laser effect and film homogeneity issues were required to be overcome. Two dimensional scan method was adapted to reduce measurement uncertainty since Raman cross-section is very sensitive to atomic bonds concentration and a large number of point measurements would guarantee reliable data group. The concentration in carbon double and triple bonds of radiochromic films would change by polymerization process. Thus, two dimensional analysis based on Raman mapping provides more reliable data in light of polymerization quantity due to radiation ionization than optical scanning. Its high spatial resolution (fifty micrometers) and low dose sensitivity (10 cGy) were demonstrated as a potential dosimeter. Raman analysis is expected as more precise analysis for micro-dosimetry in the future

  8. MCNP modelling of vaginal and uterine applicators used in intracavitary brachytherapy and comparison with radiochromic film measurements

    Science.gov (United States)

    Ceccolini, E.; Gerardy, I.; Ródenas, J.; van Dycke, M.; Gallardo, S.; Mostacci, D.

    Brachytherapy is an advanced cancer treatment that is minimally invasive, minimising radiation exposure to the surrounding healthy tissues. Microselectron© Nucletron devices with 192Ir source can be used for gynaecological brachytherapy, in patients with vaginal or uterine cancer. Measurements of isodose curves have been performed in a PMMA phantom and compared with Monte Carlo calculations and TPS (Plato software of Nucletron BPS 14.2) evaluation. The isodose measurements have been performed with radiochromic films (Gafchromic EBT©). The dose matrix has been obtained after digitalisation and use of a dose calibration curve obtained with a 6 MV photon beam provided by a medical linear accelerator. A comparison between the calculated and the measured matrix has been performed. The calculated dose matrix is obtained with a simulation using the MCNP5 Monte Carlo code (F4MESH tally).

  9. Technical Note: Magnetic field effects on Gafchromic-film response in MR-IGRT.

    Science.gov (United States)

    Reynoso, Francisco J; Curcuru, Austen; Green, Olga; Mutic, Sasa; Das, Indra J; Santanam, Lakshmi

    2016-12-01

    Magnetokinetic changes may affect crystal orientation and polymerization within the active layer of radiochromic film (RCF). This effect is investigated in a magnetic resonance image-guided radiotherapy unit within the context of film dosimetry. Gafchromic EBT2 RCF was irradiated in a 30 × 30 × 30 cm 3 solid water phantom using a Co-60 MRI guided radiotherapy system (B = 0.35 T) under normal operating conditions, and under the exact conditions and setup without a magnetic field. Fifteen 20.3 × 25.4 cm 2 EBT2 film sheets were placed at three different depths (d = 0.5, 5, and 10 cm) using five different treatment plans. The plans were computed using the MRIdian (ViewRay, Inc.) treatment planning system to deliver doses between 0 and 17.6 Gy. Films were analyzed before and after irradiation to obtain the net optical density (netOD) for each color channel separately. Scanning electron microscope (SEM) images were obtained to compare the active layer of selected samples. The results indicated that the red channel netOD decreased between 0.013 and 0.123 (average of 0.060 ± 0.033) for doses above 2.8 Gy, with a linear increase in this effect for higher doses. Green channel netOD showed similar results with a decrease between 0.012 and 0.105 (average of 0.041 ± 0.027) for doses above 3.5 Gy. The blue channel showed the weakest effect with a netOD decrease between 0.013 and 0.029 (average of 0.020 ± 0.006) for doses above 8.0 Gy. SEM images show changes in crystal orientation within active layer in RCF exposed in a magnetic field. The presence of a magnetic field affects crystal orientation and polymerization during irradiation, where netOD decreased by an average of 8.7%, 8.0%, and 4.3% in the red, green, and blue channels, respectively. The under response was dependent on dose and differed by up to 15% at 17.6 Gy.

  10. A new transmission methodology for quality assurance in radiotherapy based on radiochromic film measurements

    Science.gov (United States)

    do Amaral, Leonardo L.; Pavoni, Juliana F.; Sampaio, Francisco; Netto, Thomaz Ghilardi

    2015-01-01

    Despite individual quality assurance (QA) being recommended for complex techniques in radiotherapy (RT) treatment, the possibility of errors in dose delivery during therapeutic application has been verified. Therefore, it is fundamentally important to conduct in vivo QA during treatment. This work presents an in vivo transmission quality control methodology, using radiochromic film (RCF) coupled to the linear accelerator (linac) accessory holder. This QA methodology compares the dose distribution measured by the film in the linac accessory holder with the dose distribution expected by the treatment planning software. The calculated dose distribution is obtained in the coronal and central plane of a phantom with the same dimensions of the acrylic support used for positioning the film but in a source‐to‐detector distance (SDD) of 100 cm, as a result of transferring the IMRT plan in question with all the fields positioned with the gantry vertically, that is, perpendicular to the phantom. To validate this procedure, first of all a Monte Carlo simulation using PENELOPE code was done to evaluate the differences between the dose distributions measured by the film in a SDD of 56.8 cm and 100 cm. After that, several simple dose distribution tests were evaluated using the proposed methodology, and finally a study using IMRT treatments was done. In the Monte Carlo simulation, the mean percentage of points approved in the gamma function comparing the dose distribution acquired in the two SDDs were 99.92%±0.14%. In the simple dose distribution tests, the mean percentage of points approved in the gamma function were 99.85%±0.26% and the mean percentage differences in the normalization point doses were −1.41%. The transmission methodology was approved in 24 of 25 IMRT test irradiations. Based on these results, it can be concluded that the proposed methodology using RCFs can be applied for in vivo QA in RT treatments. PACS number: 87.55.Qr, 87.55.km, 87.55.N‐ PMID

  11. Relative dosimetry by Ebt-3

    International Nuclear Information System (INIS)

    De Leon A, M. A.; Rivera M, T.; Hernandez O, J. O.

    2015-10-01

    In the present work relative dosimetry in two linear accelerator for radiation therapy was studied. Both Varian Oncology systems named Varian Clinac 2100-Cd and MLC Varian Clinac i X were used. Gaf Chromic Ebt-3 film was used. Measurements have been performed in a water equivalent phantom, using 6 MV and 18 MV photon beams on both Linacs. Both calibration and Electron irradiations were carried out with the ionization chamber placed at the isocenter, below a stack of solid water slabs, at the depth of dose maximum (D max), with a Source-to-Surface Distance (SSD) of 100 cm and a field size of 10 cm x 10 cm. Calibration and dosimetric measurements photons were carried out under IAEA-TRS 398 protocol. Results of relative dosimetry in the present work are discussed. (Author)

  12. Gamma index evaluation of IMRT technique using gafchromic film EBT3 for homogeneous and inhomogeneous material

    Science.gov (United States)

    Nisauf, T. A.; Wibowo, W. E.; Pawiro, S. A.

    2017-07-01

    This study was done to evaluate the gamma index for registering between the planar of dose planning and the measurement of EBT film. The treatment plan was simulated for 5 patients using Fan Beam Computerized Tomography (FBCT) modality, Philips Pinnacle planning system, 6 MV photon energy, 50 segments IMRT technique, and calculation grid resolution (CGR) of 0.2 cm. Gamma Index (GI) evaluation was done with criteria of dose difference (DD) of 2 %, dose to agreement (DTA) of 2 mm and dose difference (DD) of 5 % DTA of 3 mm, SAD 100 cm, depth of 5 cm and 10 cm of the phantom. The result shows that GI for homogeneous material is greater than for inhomogeneous material with discrepancy to previous work is about 1.98 % for homogeneous material (depth 5 cm) and 2.05 % (depth 10 cm) while it was found of 2.98 % for inhomogeneous material (equivalent depth 5 cm) and 4.59 % (equivalent depth 10 cm).

  13. TH-CD-BRA-06: Magnetic Field Effects On Gafchromic-Film Response in MR-IGRT

    International Nuclear Information System (INIS)

    Reynoso, F; Curcuru, A; Green, O; Mutic, S; Santanam, L; Das, I

    2016-01-01

    Purpose: To investigate the effects of magnetic fields in radiochromic films (RCF). Magnetokinetic changes may affect crystal orientation and polymerization within active layer of RCF, these effects are investigated in Magnetic Resonance Image-guided Radiotherapy (MR-IGRT). Methods: Gafchromic EBT2 RCF were irradiated in a 30×30×30 cm 3 solid water phantom using a ViewRay MRIdian Co-60 MRI guided radiotherapy system (B=0.35 T). Fifteen 20.3×25.4 cm 2 EBT2 film sheets were placed at three different depths (d=0.5, 5 and 10 cm) and irradiated using 5 different treatment plans. The plans were computed using the MRIdian treatment planning system to deliver 2, 4, 6, 8, and 10 Gy at a depth of 10 cm. The films were scanned using an Epson Expression 10000 XL flat-bed document scanner in transmission mode. Films were processed before and after irradiation to obtain a net optical density (netOD) for each color channel separately. Scanning electron microscope (SEM) images were obtained to compare the active layer of selected samples. Results: The results show the red channel netOD decreases between 1.3–12.3 % (average of 5.95 %) for doses above 2.8 Gy, with a linear increase in this effect for higher doses. Green channel netOD showed similar results with a decrease between 1.2–10.5 % (average of 4.09 %) for doses above 3.5 Gy. Blue channel showed the weakest effect between 1.3–2.9 % (average of 1.94 %) for doses above 8.0 Gy. SEM images show changes in crystal orientation within active layer in RCF exposed in a magnetic field. Conclusion: The presence of a magnetic field affects crystal orientation and polymerization during irradiation, decreasing netOD by an average of 5.95 % in the red channel. The under response is dependent on dose and differs by up to 12.3 % at 17.6 Gy. The results show that magnetokinetic effects should be carefully considered in MR-IGRT.

  14. A definitive method for dosimetry with radiochromic movie; Un metodo definitivo para la dosimetria con pelicula radiocromica

    Energy Technology Data Exchange (ETDEWEB)

    Miras del Rio, H.; Arrans Lara, R.

    2013-07-01

    Despite the undoubted benefits of radiochromic film as a two-dimensional radiation detector, the manufacturing process and characteristics of scanners make to present some inherent difficulties that result that its use is complex, tedious, and if not performed so appropriate, imprecise. A protocol that simplifies the calibration process while reducing the number of films used for this purpose is proposed systematically corrects several difficulties, both inherent in the manufacture of the film as the reader and, what is more important achieves levels comparable to those of other more widespread use detectors accuracy. (Author)

  15. SU-F-J-151: Evaluation of a Magnetic Resonance Image Gated Radiotherapy System Using a Motion Phantom and Radiochromic Film

    Energy Technology Data Exchange (ETDEWEB)

    Lamb, J; Ginn, J; O’Connell, D; Thomas, D; Agazaryan, N; Cao, M; Yang, Y; Low, D [UCLA, Los Angeles, CA (United States)

    2016-06-15

    Purpose: Magnetic resonance image (MRI) guided radiotherapy enables gating directly on target position for soft-tissue targets in the lung and abdomen. We present a dosimetric evaluation of a commercially-available FDA-approved MRI-guided radiotherapy system’s gating performance using a MRI-compatible respiratory motion phantom and radiochromic film. Methods: The MRI-compatible phantom was capable of one-dimensional motion. The phantom consisted of a target rod containing high-contrast target inserts which moved inside a body structure containing background contrast material. The target rod was equipped with a radiochromic film insert. Treatment plans were generated for a 3 cm diameter spherical target, and delivered to the phantom at rest and in motion with and without gating. Both sinusoidal and actual tumor trajectories (two free-breathing trajectories and one repeated-breath hold) were used. Gamma comparison at 5%/3mm was used to measure fidelity to the static target dose distribution. Results: Without gating, gamma pass rates were 24–47% depending on motion trajectory. Using our clinical standard of repeated breath holds and a gating window of 3 mm with 10% of the target allowed outside the gating boundary, the gamma pass rate was 99.6%. Relaxing the gating window to 5 mm resulted in gamma pass rate of 98.6% with repeated breath holds. For all motion trajectories gated with 3 mm margin and 10% allowed out, gamma pass rates were between 64–100% (mean:87.5%). For a 5 mm margin and 10% allowed out, gamma pass rates were between 57–98% (mean: 82.49%), significantly lower than for 3 mm by paired t-test (p=0.01). Conclusion: We validated the performance of respiratory gating based on real-time cine MRI images with the only FDA-approved MRI-guided radiotherapy system. Our results suggest that repeated breath hold gating should be used when possible for best accuracy. A 3 mm gating margin is statistically significantly more accurate than a 5 mm gating margin.

  16. Simple annulus power balance in EBT-I

    International Nuclear Information System (INIS)

    Borowski, S.K.; Uckan, N.A.; Jaeger, E.F.; Kammash, T.

    1979-09-01

    An essential feature of the ELMO Bumpy Torus (EBT) concept is the presence of a relativistic electron annulus in each of the toroidal mirror sectors. These high beta annuli are formed and sustained by microwave heating and are of sufficient density and temperature that diamagnetic currents produce the necessary minimum in the magnetic field required for MHD stability of the toroidal core plasma. Because electron rings play an important role in confinement characteristics and performance of EBT, the trade-off between the quality of the confinement afforded by the rings and the power required to sustain the rings represents an important problem in a fusion reactor. Theoretical estimates of the microwave power required to sustain the annulus are found to be within a factor of 2 of the experimentally determined value. Scaling projections that are shown for both EBT-I and EBT-S enable one to examine the sensitivity of the annulus electron temperature as a function of core plasma density for various microwave power levels. The results are found to be sensitive to the details of the hot electron distribution function as well as geometric and scaling parameters. Improvements to the model are under way in order to increase its capability and accuracy in assessing the overall power balance

  17. SU-F-T-173: One-Scan Protocol: Verifying the Delivery of Spot-Scanning Proton Beam

    Energy Technology Data Exchange (ETDEWEB)

    Chan, M; Li, J [Memorial Sloan-Kettering Cancer Center, Basking Ridge, NJ (United States); Chen, C; Mah, D [Procure Treatment Center, Somerset, NJ (United States); Tang, X [Memorial Sloan Kettering Cancer Center, West Harrison, NY (United States); Li, X [Memorial Sloan Kettering Cancer Center, Rockville Centre, NY (United States); Tang, G [Memorial Sloan Kettering Cancer Center, New York, NY (United States)

    2016-06-15

    Purpose: Radiochromic film for spot-scanning QA provides high spatial resolution and efficiency gains from one-shot irradiation for multiple depths. However, calibration can be a tedious procedure which may limit widespread use. Moreover, since there may be an energy dependence, which manifests as a depth dependence, this may require additional measurements for each patient. We present a one-scan protocol to simplify the procedure. Methods: We performed the calibration using an EBT3 film at depths of 18, 20, 24cm of Plastic Water exposed by a 6-level step-wedge plan on a Proteus Plus proton system (IBA, Belgium). The calibration doses ranged 65–250 cGy(RBE) for proton energies of 170–200MeV. A clinical prostate+nodes plan was used for validation. The planar doses at selected depths were measured with EBT3 films and analyzed using one-scan protocol (one-scan digitization of QA film and at least one film exposed to known dose). The Gamma passing rates, dose-difference maps, and profiles of 2D planar doses measured with EBT3 film, IBA MatriXX PT, versus TPS calculations were analyzed and compared. Results: The EBT3 film measurement results matched well with the TPS calculation data with an average passing rate of ∼95% for 2%/2mm and slightly lower passing rates were obtained from an ion chamber array detector. We were able to demonstrate that the use of a proton step-wedge provided clinically acceptable results and minimized variations between film-scanner orientation, inter-scan, and scanning conditions. Furthermore, it could be derived from no more than two films exposed to known doses (one could be zero) for rescaling the master calibration curve at each depth. Conclusion: The use of a proton step-wedge for calibration of EBT3 film increases efficiency. The sensitivity of the calibration to depth variations has been explored. One-scan protocol results appear to be comparable to that of the ion chamber array detector. One author has a research grant from

  18. Density characterization of radiochromic film through source axis distance (SAD) technique in linac with slab phantom for radiotherapy applications

    Science.gov (United States)

    Hariani, Yousida; Haris, Bambang

    2017-05-01

    Characterization of radiochromic film density is accomplished through Source Axis Distance (SAD) technique in a slab phantom Linac with various depths and breadths of field. Type of the film used is gafchromic RTQA2. The dose of radiation exposure of the film may cause changes in the film density. This research aims to determine the relation between the density and the dose depth through the characteristic of curves to identify the depth of the dose and particular breadth of the field as a reference for the dose of radiotherapy patients. The result shows that the higher the dose is absorbed, the darker the film will be, yet the lower the density is obtained. The dose depth is determined by measuring the amount of dose received at various depths and breadths of field using film that is placed on the slab phantom with 6 MV linac radiation and dose of 300 cGy. The variation of the depth at 1.5 cm; 4 cm; 6 cm; 8 cm; 10 cm, the field size at 4 × 4 cm2, and the dose depth at 359.7 cGy; 315.3 cGy; 281.4 cGy; 241.2 cGy; 220.5 cGy were settled. The field size 6 × 6 cm2 takes the dose depth 354.6 cGy; 314.1 cGy; 282.6 cGy; 244.5 cGy; 224.7 cGy. The field size 8 × 8 cm2 takes the dose depth 351.6 cGy; 313 cGy; 283.8 cGy; 247.2 cGy; 228 cGy. The field size 10 × 10 cm2 takes the dose depth 348.9 cGy; 342.6 cGy; 248.4 cGy; 249.6 cGy; 231 cGy.

  19. Density characterization of radiochromic film through source axis distance (SAD) technique in linac with slab phantom for radiotherapy applications

    International Nuclear Information System (INIS)

    Hariani, Yousida; Haris, Bambang

    2017-01-01

    Characterization of radiochromic film density is accomplished through Source Axis Distance (SAD) technique in a slab phantom Linac with various depths and breadths of field. Type of the film used is gafchromic RTQA2. The dose of radiation exposure of the film may cause changes in the film density. This research aims to determine the relation between the density and the dose depth through the characteristic of curves to identify the depth of the dose and particular breadth of the field as a reference for the dose of radiotherapy patients. The result shows that the higher the dose is absorbed, the darker the film will be, yet the lower the density is obtained. The dose depth is determined by measuring the amount of dose received at various depths and breadths of field using film that is placed on the slab phantom with 6 MV linac radiation and dose of 300 cGy. The variation of the depth at 1.5 cm; 4 cm; 6 cm; 8 cm; 10 cm, the field size at 4 × 4 cm 2 , and the dose depth at 359.7 cGy; 315.3 cGy; 281.4 cGy; 241.2 cGy; 220.5 cGy were settled. The field size 6 × 6 cm 2 takes the dose depth 354.6 cGy; 314.1 cGy; 282.6 cGy; 244.5 cGy; 224.7 cGy. The field size 8 × 8 cm 2 takes the dose depth 351.6 cGy; 313 cGy; 283.8 cGy; 247.2 cGy; 228 cGy. The field size 10 × 10 cm 2 takes the dose depth 348.9 cGy; 342.6 cGy; 248.4 cGy; 249.6 cGy; 231 cGy. (paper)

  20. Verification of an algorithm of cono collapsed through the IAEA TECDOC 1583 protocol and dosimetry with radiochromic films; Verificacion de un algoritmo de cono colapso mediante en protocolo IAEA TECDOC 1583 y dosimetria con peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Martin-Viera Cueto, J. A.; Benitez Villegas, E. M.; Bodineau Gil, C.; Parra Osorio, V.; Garcia Pareja, S.; Casado Villalon, F. J.

    2013-07-01

    The objective of this study is to verify the characterization of the collapsed cone algorithm of an SP using this Protocol. In addition, given that it only offers details of dose values measured at discrete points, measures are complemented by a gamma test distributions 2D of doses in different cases using film radiochromic. (Author)

  1. EBTS: DESIGN AND EXPERIMENTAL STUDY

    International Nuclear Information System (INIS)

    PIKIN, A.; ALESSI, J.; BEEBE, E.; KPONOU, A.; PRELEC, K.; KUZNETSOV, G.; TIUNOV, M.

    2000-01-01

    Experimental study of the BNL Electron Beam Test Stand (EBTS), which is a prototype of the Relativistic Heavy Ion Collider (RHIC) Electron Beam Ion Source (EBIS), is currently underway. The basic physics and engineering aspects of a high current EBIS implemented in EBTS are outlined and construction of its main systems is presented. Efficient transmission of a 10 A electron beam through the ion trap has been achieved. Experimental results on generation of multiply charged ions with both continuous gas and external ion injection confirm stable operation of the ion trap

  2. Establishing an individual dosing system for patients undergoing interventional transcatheter arterial embolization: Radiochromic film and Monte Carlo simulation

    International Nuclear Information System (INIS)

    Tsai, Hui-Yu; Lai, Pei-Ling; Li, Yang-Ying; Tyan, Yeu-Sheng

    2011-01-01

    Less invasive imaging-guided vascular interventions with fluoroscopy and digital subtraction angiography have recently become widespread and have been successfully used for treating various diseases. However, interventional fluoroscopy procedures may present deterministic and stochastic radiation risks. The International Commission on Radiological Protection (ICRP) and the Food and Drug Administration have requested identifying procedures that may involve patient doses greater than the recommended thresholds. In this study, radiochromic dosimetric media, known as self-developing films, and measurement-based Monte Carlo simulations were used to establish an interventional radiology dosing system for individual patients undergoing interventional transcatheter arterial embolization. The peak skin dose, evaluated from the entrance surface dose distribution, was 21% less than the cumulated dose reported from the console. A 3D dose map incorporated into CT images was established. The organ doses and effective doses for individual patients were evaluated using this dosing system. This system could be applied very well to other fluoroscopic or interventional procedures for patient dose management.

  3. Physics and engineering aspects of the EBT reactor

    International Nuclear Information System (INIS)

    Uckan, N.A.; Bettis, E.S.; Hedrick, C.L.; Santoro, R.T.; Watts, H.L.; Yeh, H.T.

    1977-01-01

    The ELMO Bumpy Torus (EBT) reactor has the advantage of high-β, steady-state operation. The first reactor study based on the EBT confinement concept was initiated in 1976. It provided the required starting point for continued assessment of the validity of the concept. A new design based on the present physics understanding, practical design approaches, and present and near-term technologies has been established. One of the important factors in an EBT reactor is the large aspect ratio (large toroidal major radius as well). This leads to a power plant with a comparatively large total energy output, usually in the range of 2000-6000 MW(th) for a conventional neutron wall loading of 1-2 MW/m 2 (the high value of β in an EBT device provides a net cost per unit energy roughly equal to or somewhat less than that for a Tokamak system). The large aspect ratio also provides very simple engineering and design requirements because of good access and small force loading asymmetries. Another important factor is the steady-state operation. In an EBT system, less power handling, energy storage, and filtering equipment will be needed. An EBT reactor is less likely to be subject to thermal and mechanical fatigue than reactors with large pulsed magnetic fields and short bursts of fusion power. The details of the key design elements and critical scientific and technology factors which are substantially different from other fusion reactor approaches are described

  4. Space potential profiles in ELMO Bumpy Torus (EBT) experiment

    International Nuclear Information System (INIS)

    Bieniosek, F.M.; Connor, K.A.

    1983-01-01

    Spatially resolved measurements of the electric space potential in the ELMO Bumpy Torus (EBT) have been made by a heavy ion beam probe. The EBT-I device is characterized by positive potentials in the surface plasma the order of 100 V and by a nearly symmetric potential well in the core plasma of up to 300 V with respect to the surface potential. The EBT-S device has a similar potential structure with well depth and peak potential similar to or greater than that of EBT-I. Peak potential and well depth increase as the edge gas pressure is lowered and as the microwave power is increased. The potential structure is strongly linked to the specific heating geometry. The ambipolar electric field is large enough generally to dominate the core electron neoclassical diffusion. The potential profile is approximately parabolic in the core, which is shown to be a natural consequence of the spatially uniform plasma source function

  5. In vivo dosimetry in intraoperative electron radiotherapy. microMOSFETs, radiochromic films and a general-purpose linac

    Energy Technology Data Exchange (ETDEWEB)

    Lopez-Tarjuelo, Juan; Marco-Blancas, Noelia de; Santos-Serra, Agustin; Quiros-Higueras, Juan David [Consorcio Hospitalario Provincial de Castellon, Servicio de Radiofisica y Proteccion Radiologica, Castellon de la Plana (Spain); Bouche-Babiloni, Ana; Morillo-Macias, Virginia; Ferrer-Albiach, Carlos [Consorcio Hospitalario Provincial de Castellon, Servicio de Oncologia Radioterapica, Castellon de la Plana (Spain)

    2014-11-15

    In vivo dosimetry is desirable for the verification, recording, and eventual correction of treatment in intraoperative electron radiotherapy (IOERT). Our aim is to share our experience of metal oxide semiconductor field-effect transistors (MOSFETs) and radiochromic films with patients undergoing IOERT using a general-purpose linac. We used MOSFETs inserted into sterile bronchus catheters and radiochromic films that were cut, digitized, and sterilized by means of gas plasma. In all, 59 measurements were taken from 27 patients involving 15 primary tumors (seven breast and eight non-breast tumors) and 12 relapses. Data were subjected to an outliers' analysis and classified according to their compatibility with the relevant doses. Associations were sought regarding the type of detector, breast and non-breast irradiation, and the radiation oncologist's assessment of the difficulty of detector placement. At the same time, 19 measurements were carried out at the tumor bed with both detectors. MOSFET measurements (D = 93.5 %, s{sub D} = 6.5 %) were not significantly shifted from film measurements (D = 96.0 %, s{sub D} = 5.5 %; p = 0.109), and no associations were found (p = 0.526, p = 0.295, and p = 0.501, respectively). As regards measurements performed at the tumor bed with both detectors, MOSFET measurements (D = 95.0 %, s{sub D} = 5.4 %) were not significantly shifted from film measurements (D = 96.4 %, s{sub D} = 5.0 %; p = 0.363). In vivo dosimetry can produce satisfactory results at every studied location with a general-purpose linac. Detector choice should depend on user factors, not on the detector performance itself. Surgical team collaboration is crucial to success. (orig.) [German] Die In-vivo-Dosimetrie ist wuenschenswert fuer die Ueberpruefung, Registrierung und die eventuelle Korrektur der Behandlungen in der IOERT (''Intraoperative Electron Radiation Therapy''). Unser Ziel ist die Veroeffentlichung unserer Erfahrungen beim

  6. Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film.

    Science.gov (United States)

    Sonier, Marcus; Wronski, Matt; Yeboah, Collins

    2015-03-08

    Lens dose is a concern during the treatment of facial lesions with anterior electron beams. Lead shielding is routinely employed to reduce lens dose and minimize late complications. The purpose of this work is twofold: 1) to measure dose pro-files under large-area lead shielding at the lens depth for clinical electron energies via film dosimetry; and 2) to assess the accuracy of the Pinnacle treatment planning system in calculating doses under lead shields. First, to simulate the clinical geometry, EBT3 film and 4 cm wide lead shields were incorporated into a Solid Water phantom. With the lead shield inside the phantom, the film was positioned at a depth of 0.7 cm below the lead, while a variable thickness of solid water, simulating bolus, was placed on top. This geometry was reproduced in Pinnacle to calculate dose profiles using the pencil beam electron algorithm. The measured and calculated dose profiles were normalized to the central-axis dose maximum in a homogeneous phantom with no lead shielding. The resulting measured profiles, functions of bolus thickness and incident electron energy, can be used to estimate the lens dose under various clinical scenarios. These profiles showed a minimum lead margin of 0.5 cm beyond the lens boundary is required to shield the lens to ≤ 10% of the dose maximum. Comparisons with Pinnacle showed a consistent overestimation of dose under the lead shield with discrepancies of ~ 25% occur-ring near the shield edge. This discrepancy was found to increase with electron energy and bolus thickness and decrease with distance from the lead edge. Thus, the Pinnacle electron algorithm is not recommended for estimating lens dose in this situation. The film measurements, however, allow for a reasonable estimate of lens dose from electron beams and for clinicians to assess the lead margin required to reduce the lens dose to an acceptable level.

  7. SU-F-T-76: Total Skin Electron Therapy: An-End-To-End Examination of the Absolute Dosimetry with a Rando Phantom

    Energy Technology Data Exchange (ETDEWEB)

    Cui, G; Ha, J; Zhou, S; Cui, J; Shiu, A [University Southern California, Los Angeles, CA (United States)

    2016-06-15

    Purpose: To examine and validate the absolute dose for total skin electron therapy (TSET) through an end-to-end test with a Rando phantom using optically stimulated luminescent dosimeters (OSLDs) and EBT3 radiochromic films. Methods: A Varian Trilogy linear accelerator equipped with the special procedure 6 MeV HDTSe- was used to perform TSET irradiations using a modified Stanford 6-dual-field technique. The absolute dose was calibrated using a Markus ion chamber at a reference depth of 1.3cm at 100 cm SSD with a field size of 36 × 36 cm at the isocenter in solid water slabs. The absolute dose was cross validated by a farmer ion chamber. Then the dose rate in the unit of cGy/Mu was calibrated using the Markus chamber at the treatment position. OSLDs were used to independently verify the dose using the calibrated dose rate. Finally, a patient treatment plan (200 cGy/cycle) was delivered in the QA mode to a Rando phantom, which had 16 pairs of OSLDs and EBT3 films taped onto its surface at different anatomical positions. The doses recorded were read out to validate the absolute dosimetry for TSET. Results: The OSLD measurements were within 7% agreement with the planned dose except the shoulder areas, where the doses recorded were 23% lower on average than those of the planned. The EBT3 film measurements were within 10% agreement with the planned dose except the shoulder and the scalp vertex areas, where the respective doses recorded were 18% and 14% lower on average than those of the planned. The OSLDs gave more consistent dose measurements than those of the EBT3 films. Conclusion: The absolute dosimetry for TSET was validated by an end-to-end test with a Rando phantom using the OSLDs and EBT3 films. The beam calibration and monitor unit calculations were confirmed.

  8. Study of film dosimetry for radiotherapy with Gafchromic-RTQ plates

    International Nuclear Information System (INIS)

    Diaz Moreno, Rogelio Manuel; Lara Mas, Elier; Alfonso Laguardia, Rodolfo

    2009-01-01

    Film dosimetry allows quality control processes (CC) for advanced radiotherapy treatments, not achievable with other types of systems dosimetry, as is the determination of two-dimensional dose distribution provided with the planned treatment in selected planes. The aim of this work was to establish the possibilities of making this type of CC with the means available in the INOR. Plates were used radiochromic Gafchromic-RTQ, for quality control, which irradiated with Elekta Precise linear accelerator, according to the test planning developed in the treatment planning system Precise Plan. Were used as image processing software the Mephysto mc2, PTW, and routines scheduled at home on Matlab. Was prepared calibration curve Dose-response for these plates, and applied this calibration curve at other boards with known radiation dose to estimate proximity of the dose obtained through calibration. Other tests were performed to determine the conditions of repeatability and optimal parameters of the process. Conditions were established that are obtained more reliable , the which are lower than those reported Gafchromic-EBT plates, especially designed quantitative dosimetric purposes, but in certain ranges allow evaluate the of a plan with an acceptable degree of approximation. (author)

  9. Small Radiation Beam Dosimetry for Radiosurgery of Trigeminal Neuralgia: One Case Analysis

    International Nuclear Information System (INIS)

    Garcia-Garduno, O. A.; Larraga-Gutierrez, J. M.; Rodriguez-Villafuerte, M.; Martinez-Davalos, A.; Moreno-Jimenez, S.; Suarez-Campos, J. J.; Celis, M. A.

    2008-01-01

    The use of small radiation beams for trigeminal neuralgia (TN) treatment requires high precision and accuracy in dose distribution calculations and delivery. Special attention must be kept on the type of detector to be used. In this work, the use of GafChromic EBT registered radiochromic and X-OMAT V2 radiographic films for small radiation beam characterization is reported. The dosimetric information provided by the films (total output factors, tissue maximum ratios and off axis ratios) is compared against measurements with a shielded solid state (diode) reference detector. The film dosimetry was used for dose distribution calculations for the treatment of trigeminal neuralgia radiosurgery. Comparison of the isodose curves shows that the dosimetry produced with the X-OMAT radiographic film overestimates the dose distributions in the penumbra region

  10. How flatbed scanners upset accurate film dosimetry

    Science.gov (United States)

    van Battum, L. J.; Huizenga, H.; Verdaasdonk, R. M.; Heukelom, S.

    2016-01-01

    Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2-2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un)irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner’s transmission mode, with red-green-blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner’s optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the film.

  11. How flatbed scanners upset accurate film dosimetry

    International Nuclear Information System (INIS)

    Van Battum, L J; Verdaasdonk, R M; Heukelom, S; Huizenga, H

    2016-01-01

    Film is an excellent dosimeter for verification of dose distributions due to its high spatial resolution. Irradiated film can be digitized with low-cost, transmission, flatbed scanners. However, a disadvantage is their lateral scan effect (LSE): a scanner readout change over its lateral scan axis. Although anisotropic light scattering was presented as the origin of the LSE, this paper presents an alternative cause. Hereto, LSE for two flatbed scanners (Epson 1680 Expression Pro and Epson 10000XL), and Gafchromic film (EBT, EBT2, EBT3) was investigated, focused on three effects: cross talk, optical path length and polarization. Cross talk was examined using triangular sheets of various optical densities. The optical path length effect was studied using absorptive and reflective neutral density filters with well-defined optical characteristics (OD range 0.2–2.0). Linear polarizer sheets were used to investigate light polarization on the CCD signal in absence and presence of (un)irradiated Gafchromic film. Film dose values ranged between 0.2 to 9 Gy, i.e. an optical density range between 0.25 to 1.1. Measurements were performed in the scanner’s transmission mode, with red–green–blue channels. LSE was found to depend on scanner construction and film type. Its magnitude depends on dose: for 9 Gy increasing up to 14% at maximum lateral position. Cross talk was only significant in high contrast regions, up to 2% for very small fields. The optical path length effect introduced by film on the scanner causes 3% for pixels in the extreme lateral position. Light polarization due to film and the scanner’s optical mirror system is the main contributor, different in magnitude for the red, green and blue channel. We concluded that any Gafchromic EBT type film scanned with a flatbed scanner will face these optical effects. Accurate dosimetry requires correction of LSE, therefore, determination of the LSE per color channel and dose delivered to the film. (paper)

  12. Estimation of dose enhancement to soft tissue due to backscatter radiation near metal interfaces during head and neck radiothearpy - A phantom dosimetric study with radiochromic film

    Directory of Open Access Journals (Sweden)

    Rajesh Ashok Kinhikar

    2014-01-01

    Full Text Available The objective of this study was to investigate the dose enhancement to soft tissue due to backscatter radiation near metal interfaces during head and neck radiotherapy. The influence of titanium-mandibular plate with the screws on radiation dose was tested on four real bones from mandible with the metal and screws fixed. Radiochromic films were used for dosimetry. The bone and metal were inserted through the film at the center symmetrically. This was then placed in a small jig (7 cm × 7 cm × 10 cm to hold the film vertically straight. The polymer granules (tissue-equivalent were placed around the film for homogeneous scatter medium. The film was irradiated with 6 MV X-rays for 200 monitor units in Trilogy linear accelerator for 10 cm × 10 cm field size with source to axis distance of 100 cm at 5 cm. A single film was also irradiated without any bone and metal interface for reference data. The absolute dose and the vertical dose profile were measured from the film. There was 10% dose enhancement due to the backscatter radiation just adjacent to the metal-bone interface for all the materials. The extent of the backscatter effect was up to 4 mm. There is significant higher dose enhancement in the soft tissue/skin due to the backscatter radiation from the metallic components in the treatment region.

  13. Analysis of energy dependence of the film radiochromic XRQA2 for seven spectra of X-ray standard in diagnostic radiology; Analisis de la dependencia energetica de la pelicula radiogromica XRQA2 para siete espectros de rayos X habituales en radiodiagnostico

    Energy Technology Data Exchange (ETDEWEB)

    Vicent Granado, D.; Gutierrez Ramos, S. M.; Sanchez Carmona, G.; Santos Rubio, A. J.; Herrador Cordoba, M.

    2013-07-01

    The response of the radiochromic film XRQA2 depends on the peak voltage, as well as filtration and measurement with or without backscatter. The common method of using a calibration from the kerma in air to measure the dose at the entrance of patients is not the most advisable for this film in this dose range. Whenever possible should be the calibration with the spectrum of X rays for which will be the measurement of absorbed dose. (Author)

  14. Diagnostic value of electron-beam computed tomography (EBT). I. cardiac applications

    International Nuclear Information System (INIS)

    Enzweiler, C.N.H.; Lembcke, A.; Rogalla, P.; Taupitz, M.; Wiese, T.H.; Hammm, B.; Becker, C.R.; Reiser, M.F.; Felix, R.; Knollmann, F.D.; Georgi, M.; Weisser, G.; Lehmann, K.J.

    2004-01-01

    Electron beam tomography (EBT) directly competes with other non-invasive imaging modalities, such as multislice computed tomography, magnetic resonance imaging, and echocardiography, in the diagnostic assessment of cardiac diseases. EBT is the gold standard for the detection and quantification of coronary calcium as a preclinical sign of coronary artery disease (CAD). Its standardized examination protocols and the broad experience with this method favor EBT. First results with multislice CT indicate that this new technology may be equivalent to EBT for coronary calcium studies. The principal value of CT-based coronary calcium measurements continues to be an issue of controversy amongst radiologists and cardiologists due to lack of prospective randomized trials. Coronary angiography with EBT is characterized by a high negative predictive value and, in addition, may be indicated in some patients with manifest CAD. It remains to be shown whether coronary angiography with multislice CT is reliable and accurate enough to be introduced into the routine work-up, to replace some of the many strictly diagnostic coronary catheterizations in Germany and elsewhere. Assessment of coronary stent patency with EBT is associated with several problems and in our opinion cannot be advocated as a routine procedure. EBT may be recommended for the evaluation of coronary bypasses to look for bypass occlusions and significant stenoses, which, however, can be equally well achieved with multislice CT. Quantification of myocardial perfusion with EBT could not replace MRI or other modalities in this field. EBT has proven to be accurate, reliable and in some instances equivalent to MRI, which is the gold standard for the quantitative and qualitative evaluation of cardiac function. Some disadvantages not the least of which is the limited distribution of electron beam scanners, favor MRI for functional assessment of the heart. (orig.) [de

  15. Economic analysis of EBT reactor

    International Nuclear Information System (INIS)

    Woo, J.T.; Uckan, N.A.; Lidsky, L.M.

    1977-01-01

    In order to establish the economic potential of the Elmo Bumpy Torus (EBT) reactor, two independent system-costing models have been developed. Both models predict capital costs of approximately $400/kW(th). These relatively low costs reflect the simplicity of the EBTR design. In particular, the modular nature of the individual blanket-shield segments, the low costs ''accelerator style'' containment building, high beta, and steady-state operation lead to relatively low reactor costs. A detailed cost breakdown for subsystems is analyzed. High cost and high uncertainty subsystems are identified to direct further emphasis into those areas. The calculated capital costs for the EBT reactor are compared with those costs quoted for tokamak reactors

  16. Calibration of film radiochromic EBT2 for sources of I-125 encapsulated

    International Nuclear Information System (INIS)

    Huerga Cabrerizo, C.; Luquero Llopis, N.; Torre Hernandez, I. de la; Ferrer Garcia, C.; Corredoira silva, E.; Serrada Hierro, A.

    2013-01-01

    This paper determines the calibration curve in absolute dose for sources of I-125 encapsulated to estimate its uncertainty. In order to assess energy dependence is compared with the obtained for an accelerator of 6MV calibration curve. (Author)

  17. Value of electron beam tomography (EBT). II. non-cardiac applications and radiation exposure

    International Nuclear Information System (INIS)

    Enzweiler, C.N.H.; Lembcke, A.; Rogalla, P.; Taupitz, M.; Wiese, T.H.; Hamm, B.; Becker, C.R.; Bruening, R.; Reiser, M.F.; Schoepf, U.J.; Felix, R.; Knollmann, F.D.; Georgi, M.; Weisser, G.; Lehmann, K.J.

    2004-01-01

    Electron beam tomography (EBT) has been scientifically evaluated to a much lesser degree for non-cardiac indications than for cardiac purposes. Therefore, four groups of investigators in Berlin (2), Mannheim and Muenchen, which were supported by the Deutsche Forschungsgemeinschaft (DFG), included applications outside the heart in their evaluation of EBT technology. EBT has proven useful to look for pulmonary embolism and to assess other vessels (aorta, aortic branches, and intracranial arteries). Imaging of the lung parenchyma benefits from its intrinsic high contrast and from the fast data acquisition of EBT. Limited photon efficiency, higher radiation exposure, increased noise levels and other artifacts, however markedly reduce the value of EBT for imaging of low contrast objects compared to conventional spiral CT and multislice CT (MSCT), compromising, in particular, the morphologic depiction of parenchymal abdominal organs and the brain. Consequently, scientific studies to further evaluate EBT for scanning of the brain and parenchymal abdominal organs were not pursued. Radiation exposure for non-cardiac EBT studies is up to three times higher than that for respective spiral CT studies, and in children EBT can only be advocated in select cases. Radiation exposure for the various prospectively triggered cardiac examination protocols of EBT is lower than that for conventional coronary angiography. Radiation exposure in cardiac multislice CT exceeds severalfold that of EBT, but the dose efficiency of EBT and MSCT are similar due to higher spatial resolution and less image noise of MSCT. In addition, modifications of MSCT (ECG pulsing) can further reduce radiation exposure to the level of EBT. Technical improvements of the EBT successor scanner 'e-Speed' enable faster data acquisiton at higher spatial resolution. Within comparative studies, the 'e-Speed' will have to prove its value and competitiveness, particularly in comparison with multislice CT. After profound

  18. SU-G-201-12: Investigation of Beta-Emitter 90Sr-90Y Dose Distribution Using Gafchromic EBT3 Film for Application On Conformal Skin Brachytherapy Device

    International Nuclear Information System (INIS)

    Ferreira, C; Johnson, D; Ahmad, S; Rasmussen, K; Jung, J

    2016-01-01

    Purpose: To investigate 90 Sr- 90 Y as a high dose rate (HDR) source for application in a conformal skin brachytherapy (CSBT) device. The CSBT device has been previously developed to provide patient specific treatment for small inoperable lesions and irregular surfaces. Methods: A popular beta emitter, 90 Sr- 90 Y was tested for feasibility in a CSBT device. A 1 cm diameter plaque was used to deliver dose to a solid water phantom containing EBT3 Gafchromic films arranged at the surface and perpendicular to it. Additionally, a 1 cm diameter 6 MeV electron beam was used to irradiate EBT3 film at 100 cm SSD with a 0.5 cm bolus. Films were digitized with an Epson Expression 10000 XL scanner and calibrated with a 6 MeV electron specific dose curve. Normalized percent depth doses (PDD) and dose profiles for both techniques were analyzed using ImageJ. Results: Dose distributions achieved with the 90 Sr- 90 Y sources were compared with those of external electron beam radiation therapy (EBRT). Penumbra (20%- 80%) for EBRT and 90Sr-90Y were 4.3 mm and 1.6 mm, respectively. PDD values of 50% (normalized to 2 mm) were 10.1 mm and 2.8 mm for electron and 90 Sr- 90 Y, respectively. Flatness (80% of the central beam profile) was 14.1% at a 5 mm depth for EBRT and 4.0% at surface for the 90 Sr- 90 Y. Conclusion: As expected, the PDDs of 90 Sr- 90 Y in water are shallower than that of external electron beams for the same field size. 90 Sr- 90 Y can be used in CSBT to provide patient specific treatment where shallower depth doses than that provided by electron external beams may be required: e.g. eyelids, nose, lips, ears, etc. The customizability of EBRT could be replicated by using multiple adjacent 90 Sr- 90 Y plaque placements.

  19. SU-G-201-12: Investigation of Beta-Emitter 90Sr-90Y Dose Distribution Using Gafchromic EBT3 Film for Application On Conformal Skin Brachytherapy Device

    Energy Technology Data Exchange (ETDEWEB)

    Ferreira, C; Johnson, D; Ahmad, S [University of Oklahoma Health Sciences Center, Oklahoma City, OK (United States); Rasmussen, K [University of Texas HSC SA, San Antonio, TX (United States); Jung, J [East Carolina University Greenville, NC (United States)

    2016-06-15

    Purpose: To investigate {sup 90}Sr-{sup 90}Y as a high dose rate (HDR) source for application in a conformal skin brachytherapy (CSBT) device. The CSBT device has been previously developed to provide patient specific treatment for small inoperable lesions and irregular surfaces. Methods: A popular beta emitter, {sup 90}Sr-{sup 90}Y was tested for feasibility in a CSBT device. A 1 cm diameter plaque was used to deliver dose to a solid water phantom containing EBT3 Gafchromic films arranged at the surface and perpendicular to it. Additionally, a 1 cm diameter 6 MeV electron beam was used to irradiate EBT3 film at 100 cm SSD with a 0.5 cm bolus. Films were digitized with an Epson Expression 10000 XL scanner and calibrated with a 6 MeV electron specific dose curve. Normalized percent depth doses (PDD) and dose profiles for both techniques were analyzed using ImageJ. Results: Dose distributions achieved with the {sup 90}Sr-{sup 90}Y sources were compared with those of external electron beam radiation therapy (EBRT). Penumbra (20%- 80%) for EBRT and 90Sr-90Y were 4.3 mm and 1.6 mm, respectively. PDD values of 50% (normalized to 2 mm) were 10.1 mm and 2.8 mm for electron and {sup 90}Sr-{sup 90}Y, respectively. Flatness (80% of the central beam profile) was 14.1% at a 5 mm depth for EBRT and 4.0% at surface for the {sup 90}Sr-{sup 90}Y. Conclusion: As expected, the PDDs of {sup 90}Sr-{sup 90}Y in water are shallower than that of external electron beams for the same field size. {sup 90}Sr-{sup 90}Y can be used in CSBT to provide patient specific treatment where shallower depth doses than that provided by electron external beams may be required: e.g. eyelids, nose, lips, ears, etc. The customizability of EBRT could be replicated by using multiple adjacent {sup 90}Sr-{sup 90}Y plaque placements.

  20. SU-F-T-569: Implementation of a Patient Specific QA Method Using EBT-XD for CyberKnife SRS/SBRT Plans

    Energy Technology Data Exchange (ETDEWEB)

    Zerouali, K; Aubry, J; Doucet, R [Centre hospitalier de l’Universite de Montreal, Montreal, Quebec (Canada)

    2016-06-15

    Purpose: To implement the new EBT-XD Gafchromic films for accurate dosimetric and geometric validation of stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) CyberKnife (CK) patient specific QA. Methods: Film calibration was performed using a triplechannel film analysis on an Epson 10000XL scanner. Calibration films were irradiated using a Varian Clinac 21EX flattened beam (0 to 20 Gy), to ensure sufficient dose homogeneity. Films were scanned to a resolution of 0.3 mm, 24 hours post irradiation following a well-defined protocol. A set of 12 QA was performed for several types of CK plans: trigeminal neuralgia, brain metastasis, prostate and lung tumors. A custom made insert for the CK head phantom has been manufactured to yield an accurate measured to calculated dose registration. When the high dose region was large enough, absolute dose was also measured with an ionization chamber. Dose calculation is performed using MultiPlan Ray-tracing algorithm for all cases since the phantom is mostly made from near water-equivalent plastic. Results: Good agreement (<2%) was found between the dose to the chamber and the film, when a chamber measurement was possible The average dose difference and standard deviations between film measurements and TPS calculations were respectively 1.75% and 3%. The geometric accuracy has been estimated to be <1 mm, combining robot positioning uncertainty and film registration to calculated dose. Conclusion: Patient specific QA measurements using EBT-XD films yielded a full 2D dose plane with high spatial resolution and acceptable dose accuracy. This method is particularly promising for trigeminal neuralgia plan QA, where the positioning of the spatial dose distribution is equally or more important than the absolute delivered dose to achieve clinical goals.

  1. SU-F-T-493: An Investigation Into the Feasibility of Using PipsPro Software with Film for Linac QA

    International Nuclear Information System (INIS)

    Underwood, R

    2016-01-01

    Purpose: To determine the feasibility of using radiochromic and radiographic film with Pipspro software for quality assurance of linear accelerators with no on-board imaging. Methods: The linear accelerator being used is a Varian Clinac 21EX. All IGRT is performed using the BrainLab ExacTrac system. Because of the lack of on board imaging, certain monthly and annual TG-142 quality assurance tests are more difficult to perform and analyze to a high degree of accuracy. Pipspro was not designed to be used with hard film, and to our knowledge its use with film had not been investigated. The film used will be GafChromic EBT3 film and Kodak EDR2 film, scanned with an Epson V700 scanner. The following routine tests will be attempted: MLC picket fence, light vs. radiation field coincidence, starshots, and MLC transmission. Results: The only tests that gave accurate and reliable results were the couch, gantry, and collimator starshots. Typical MV and kV images are acquired with a much higher level of contrast between the irradiated and non-irradiated areas when compared to film. Pipspro relies on this level of contrast to be able to automatically detect the fiducial points from its phantom devices, leaf edges for picket fence and transmission tests, and jaw edges for light vs. radiation field tests. Because of this, certain tests gave erroneous results and others were not able to be performed in the software at all, with either type of film. The number of monitor units delivered to the film, the experimental setup, and the scan settings was not able to rectify the problem. Conclusion: For linear accelerators with no on-board imaging, it is not recommended to use hard film with PipsPro to perform TG-142 quality assurance tests. Other software or methods should instead be investigated.

  2. SU-F-T-08: Brachytherapy Film Dosimetry in a Water Phantom for a Ring and Tandem HDR Applicator

    International Nuclear Information System (INIS)

    Lee, B; Grelewicz, Z; Kang, Z; Cutright, D; Gopalakrishnan, M; Sathiaseelan, V; Zhang, H

    2016-01-01

    Purpose: The feasibility of dose measurement using new generation EBT3 film was explored in a water phantom for a ring and tandem HDR applicator for measurements tracking mucosal dose during cervical brachytherapy. Methods: An experimental fixture was assembled to position the applicator in a water phantom. Prior to measurement, calibration curves for EBT3 film in water and in solidwater were verified. EBT3 film was placed at different known locations around the applicator in the water tank. A CT scan of the phantom with applicator was performed using clinical protocol. A typical cervical cancer treatment plan was then generated by Oncentra brachytherapy planning system. A dose of 500 cGy was prescribed to point A (2 cm, 2 cm). Locations measured by film included the outer surface of the ring, measurement point A-m (2.2 cm, 2.2 cm), and profiles extending from point A-m parallel to the tandem. Three independent measurements were conducted. The doses recorded by film were carefully analyzed and compared with values calculated by the treatment planning system. Results: Assessment of the EBT3 films indicate that the dose at point A matches the values predicted by the planning system. Dose to the point A-m was 411.5 cGy, and the outer circumferential surface dose of the ring was between 500 and 1150 cGy. It was found that from the point A-m, the dose drops 60% within 4.5 cm on the line parallel to the tandem. The measurement doses agree with the treatment planning system. Conclusion: Use of EBT3 film is feasible for in-water measurements for brachytherapy. A carefully machined apparatus will likely improve measurement accuracy. In a typical plan, our study found that the ring surface dose can be 2.5 times larger than the point A prescription dose. EBT3 film can be used to monitor mucosal dose in brachytherapy treatments.

  3. SU-F-T-08: Brachytherapy Film Dosimetry in a Water Phantom for a Ring and Tandem HDR Applicator

    Energy Technology Data Exchange (ETDEWEB)

    Lee, B; Grelewicz, Z; Kang, Z; Cutright, D; Gopalakrishnan, M; Sathiaseelan, V; Zhang, H [Northwestern Memorial Hospital, Chicago, IL (United States)

    2016-06-15

    Purpose: The feasibility of dose measurement using new generation EBT3 film was explored in a water phantom for a ring and tandem HDR applicator for measurements tracking mucosal dose during cervical brachytherapy. Methods: An experimental fixture was assembled to position the applicator in a water phantom. Prior to measurement, calibration curves for EBT3 film in water and in solidwater were verified. EBT3 film was placed at different known locations around the applicator in the water tank. A CT scan of the phantom with applicator was performed using clinical protocol. A typical cervical cancer treatment plan was then generated by Oncentra brachytherapy planning system. A dose of 500 cGy was prescribed to point A (2 cm, 2 cm). Locations measured by film included the outer surface of the ring, measurement point A-m (2.2 cm, 2.2 cm), and profiles extending from point A-m parallel to the tandem. Three independent measurements were conducted. The doses recorded by film were carefully analyzed and compared with values calculated by the treatment planning system. Results: Assessment of the EBT3 films indicate that the dose at point A matches the values predicted by the planning system. Dose to the point A-m was 411.5 cGy, and the outer circumferential surface dose of the ring was between 500 and 1150 cGy. It was found that from the point A-m, the dose drops 60% within 4.5 cm on the line parallel to the tandem. The measurement doses agree with the treatment planning system. Conclusion: Use of EBT3 film is feasible for in-water measurements for brachytherapy. A carefully machined apparatus will likely improve measurement accuracy. In a typical plan, our study found that the ring surface dose can be 2.5 times larger than the point A prescription dose. EBT3 film can be used to monitor mucosal dose in brachytherapy treatments.

  4. Evaluation of a new self-developing instant film for imaging and dosimetry

    International Nuclear Information System (INIS)

    Watanabe, Y.; Patel, G. N.; Patel, P.

    2006-01-01

    Radiation sensitive films are standard dosimetric tools in radiation therapy. Films are used for machine quality assurance (QA) and treatment planning software evaluation. With the advent of intensity modulated radiation therapy (IMRT), simple and fast imaging technology is needed for patient-specific verification of radiation fields. Conventional radiographic films are often used. Radiochromic films, e.g. Gafchromic films, were recently introduced to the market. But these films have some disadvantages. JP Laboratories have developed a prototype radiochromic film, called SIFID (self-developing, instant film for imaging and dosimetry) with superior performance such that SIFID is unaffected by ambient light for months, stable up to 90 deg. C and can be archived. SIFID is made of polymerizable diacetylene. The film develops blue colour instantly upon absorbing radiation. We evaluated the film for radiation therapy applications. Our preliminary data demonstrate its feasibility as a dosimetric tool for IMRT QA as well as for other applications. (authors)

  5. Skin Dosimetry in Breast Teletherapy on a Phantom Anthropomorphic and Anthropometric Phantom

    International Nuclear Information System (INIS)

    Batista Nogueira, Luciana; Lemos Silva, Hugo Leonardo; Donato da Silva, Sabrina; Passos Ribeiro Campos, Tarcisio

    2015-01-01

    This paper addresses the breast teletherapy dosimetry. The goal is to evaluate and compare absorbed doses in equivalent skin tissue, TE-skin, of an anthropomorphic and anthropometric breast phantom submitted to breast radiotherapy. The methodology involved the reproduction of a set of tomographic images of the phantom; the elaboration of conformational radiotherapy planning in the SOMAVISION and CadPlan (TPS) software; and the synthetic breast irradiation by parallel opposed fields in 3D conformal teletherapy at 6 MV linear accelerator Clinac-2100 C from VARIAN with prescribed dose (PD) of 180 cGy to the target volume (PTV), referent to the glandular tissue. Radiochromic films EBT2 were selected as dosimeters. Two independent calibration processes of films with solid water Gammex 457 plates and water filled box were produced. Curves of optical density (OD) versus absorbed dose were produced. Dosimeters were positioned in the external region of the breast phantom in contact with TE-skin, area of 4.0 cm 2 each. The irradiation process was prepared in duplicate to check the reproducibility of the technique. The radiochromic films were scanned and their response in RGB (Red, Green, Blue) analyzed by the ImageJ software. The optical density was obtained and converted to dose based on the calibration curves. Thus, the spatial dose distribution in the skin was reproduced. The absorbed doses measured on the radiochromic films in TE-skin showed values between upper and lower quadrants at 9 o'clock in the range of 54% of PD, between the upper and lower quadrants 3 o'clock in the range of 72% and 6 o'clock at the lower quadrant in the range of 68 % of PD. The values are ±64% (p <0.05) according to the TPS. It is concluded that the depth dose measured in solid water plates or water box reproduce equivalent dose values for both calibration processes of the radiochromic films. It was observed that the skin received doses ranging from 50% to 78% of the prescribed

  6. Development and dosimetric evaluation of radiochromic PCDA vesicle gel dosimeters

    International Nuclear Information System (INIS)

    Sun, P.; Fu, Y.C.; Hu, J.; Hao, N.; Huang, W.; Jiang, B.

    2016-01-01

    The gel dosimeter has the unique capacity in recording radiation dose distribution in three dimensions (3D), which has the specific advantages in dosimetry measurements where steep dose gradients exist, such as in intensity-modulated radiation therapy (IMRT), brachytherapy and so on. Some 3D dosimeters, such as Fricke gel dosimeters, polymer gel dosimeters, the PRESAGE plastic dosimeters and micelle gel dosimeters have appeared recently. However, there are several disadvantages of these 3D dosimeters limit their application in radiotherapy dose verification. In this study, a novel radiochromic gel dosimeter for 3D dose verification of radiotherapy was developed by dispersing nanovesicles self-assembled by 10,12-pentacosadiynoic acid (PCDA) into the tissue equivalence gel matrix. The characteristics of radiochromic PCDA vesicle gel dosimeters were evaluated. The results indicate that these radiochromic gel dosimeters have good linear dose response to X-ray irradiation in the dose range of 2–100 Gy. In addition, the radiochromic gel dosimeters breakthrough the limitations of the existing gel dosimeters such as diffusion effect, post-radiation effect, and poor forming ability. The response of the gel dosimeter does not show any dose rate dependence, energy dependence and temperature effect, and there was no obvious difference in the gel response between single and cumulative dose of fractional irradiation. Hence, the radiochromic PCDA vesicle gel dosimeters developed in this study could be generally applied to 3D dose verification in radiotherapy. - Highlights: • A novel radiochromic gel dosimeter was developed by dispersing PCDA nanovesicles into the tissue equivalence gel matrix. • This nanovesicle overcomes the dose image blurring caused by the diffusion of monomer molecules. • This nanovesicle limits the polymer chain growth, so as to reduce the post-radiation effect. • The gel matrixes possess excellent tissue equivalence and elastic strength, which

  7. Supporting material for EBT-P review presentation at DOE Germantown, 9-11 January 1980: working notes

    International Nuclear Information System (INIS)

    Sheffield, J.

    1980-01-01

    This memorandum reviews the EBT theory of Hedrick et al., and using a model of the EBT-1/S plasma simplifies the transport equations to a few relations involving the radial electric field, radial flow, diffusion coefficients, input power and measured plasma parameters. A summary is given of EBT-1/S experimental data with reference data sets and best data sets. Theory and experiment are compared and it is shown that the electron diffusion is classical within a factor of 1.5. The proposed EBT-1/S program is reviewed and it is shown how it is designed to provide information for EBT-P in a timely fashion and fill in the gaps of knowledge in the EBT-1/S studies. Finally a brief discussion is given of EBT-Q and it is shown how the EBT-P program is designed to provide information for this ignition device

  8. Matchline dosimetry in step and shoot IMRT fields: a film study

    International Nuclear Information System (INIS)

    Tangboonduangjit, P; Metcalfe, P; Butson, M; Quach, K Y; Rosenfeld, A

    2004-01-01

    The Varian millennium 120 multileaf collimator has curved leaf ends. Transmission through the leaf ends generates a small asymmetric penumbral dose effect. This design can lead to hot spots between neighbouring beam segments during step and shoot IMRT dose delivery. We have observed some matchlines with film for clinical beams optimized using the pinnacle radiotherapy treatment planning system; hence we sought to verify the optimum leaf offset required to minimize the matchline effect. An in-house program was created to control the MLC leaf banks in 2 cm steps with a 2 cm gap. The gap was varied by the following offset values from 0.0 to 0.1 cm. Two types of radiographic films (Kodak EDR and XV films) and a radiochromic film (Gafchromic MD-55-2) were used to measure the optical density maps. The films were positioned in a solid water phantom perpendicular to the beam axis and irradiated at d max using a 6 MV photon beam. An ion chamber (IC4) was used to measure point doses for normalization in a beam umbral minima position. The relative mean peak to valley dose ratios measured with no leaf offset were 1.31, 1.30 and 1.31 for the XV, EDR2 and Gafchromic films, respectively. For a 0.07 cm gap per leaf and a performance of end leaf repeatability of 0.01 cm, the central matchline was reduced to about 1.0 for all dosimeters, with two mini-peaks measured as 1.05, 1.05 and 1.08 each side of the matchline, for XV, EDR2 and Gafchromic, respectively. The average relative dose across the umbra for this offset was XO-mat V = 1.01, EDR = 1.01 and radiochromic film = 1.02, respectively. While we expected the beam penumbral tails from segment neighbours to cause overprediction of the dose in the central valley regions due to the energy response of radiographic films, by normalizing all dosimeters to an ion chamber reading in the minimum we could not observe any major shape distortion between the radiographic film and radiochromic film results. In conclusion, relative doses

  9. Selection of vacuum seals for EBT-P

    International Nuclear Information System (INIS)

    Dillow, C.F.; Adlon, G.L.; Stubblefield, V.E.

    1981-01-01

    Elmo Bumpy Torus Proof of Principle (EBT-P) is a magnetic fusion RandD facility being constructed by McDonnell Douglas Astronautics Company - St. Louis Division of Oak Ridge National Laboratory (ORNL). EBT-P is a truly unique fusion device requiring careful consideration in selecting both the primary vacuum seals on the toroidal vessel and the secondary vacuum seals in components such as vacuum pumps and and valves. The vacuum seal environment is described and the considerations in vacuum seal selection fully discussed. Methods for protecting vacuum seals in pumps and valves from the microwave environments are also presented

  10. Intensities of incident and transmitted ultraviolet-a rays through gafchromic films

    Directory of Open Access Journals (Sweden)

    Toshizo Katsuda

    2017-01-01

    Full Text Available Gafchromic films have been applied to X-ray dosimetry in diagnostic radiology. To correct nonuniformity errors in Gafchromic films, X-rays in the double-exposure technique can be replaced with ultraviolet (UV-A rays. Intensities of the incident and transmitted UV-A rays were measured. However, it is unclear whether the chemical color change of Gafchromic films affects the UV-A transmission intensity. Gafchromic EBT3 films were suitable to be used in this study because non-UV protection layers are present on both sides of the film. The film is placed between UV-A ray light-emitting diodes and a probe of a UV meter. Gafchromic EBT3 films were irradiated by UV-A rays for up to 60 min. Data for analysis were obtained in the subsequent 60 min. Images from before and after UV-A irradiation were subtracted. When using 375 nm UV-A, the mean ± standard deviation (SD of the pixel values in the subtracted image was remarkably high (11,194.15 ± 586.63. However, the UV-A transmissivity remained constant throughout the 60 min irradiation period. The mean ± SD UV-A transmission intensity was 184.48 ± 0.50 μm/cm2. Our findings demonstrate that color density changes in Gafchromic EBT3 films do not affect their UV-A transmission. Therefore, Gafchromic films were irradiated by UV-A rays as a preexposure.

  11. EFFECTS OF EBT CUSTOMER SERVICE WAIVERS ON FOOD STAMP RECIPIENTS: EXECUTIVE SUMMARY

    OpenAIRE

    Kirlin, John A.; Logan, Christopher

    2002-01-01

    Most State agencies are now using electronic benefits transfer (EBT) systems to issue food stamp benefits. To promote operational efficiency, some States have received waivers of certain rules governing EBT use. An exploratory study was conducted to ascertain the effects of these waivers on food stamp recipients. The results show that two of the waivers-those allowing recipients to select their own personal identification numbers and to receive EBT training by mail rather than in person-cause...

  12. Status of the theoretical study of microwave heating in EBT

    International Nuclear Information System (INIS)

    Batchelor, D.B.

    1978-09-01

    The basic strategy of the theoretical study of microwave heating in the ELMO Bumpy Torus (EBT) is outlined and the current status of the various aspects of the study is described. There are four broad areas which are being investigated: (1) heating and wave damping mechanisms, (2) the geometrical optics of microwave propagation in EBT, (3) reflection and mode conversion effects at regions such as cutoff and resonances where the geometrical optics approximation breaks down, and (4) nonlinear effects such as ponderamotive effects and parametric decay. Details are given of the geometrical optics code which has been developed to do ray tracing in arbitrary three dimensional (3-D) plasma equilibria. Examples are given for plasma parameters characteristic of EBT-I and EBT-II. Details are also given of the stochastic heating model currently in use with the 1-D transport code and of the linear wave damping model used in the ray tracing code. The most pressing problems of physics yet to be addressed and the directions for future work are indicated

  13. Radiochromic film containing methyl viologen for radiation dosimetry

    DEFF Research Database (Denmark)

    Lavalle, M.; Corda, U.; Fuochi, P.G.

    2007-01-01

    Poly(vinyl alcohol) (PVA) films containing methyl viologen (MV2+) that colours blue upon exposure to ionizing radiation were investigated as possible dosimeters for use in radiation processing applications. In order to find the most suitable composition of the PVA-MV2+ film, different......, humidity and temperature on the response of the PVA-MV2+ dosimeter film have been studied under laboratory conditions. We conclude that the PVA film containing MV2+ is a promising tool for the absorbed dose measurements in several industrial applications of ionizing radiations. (C) 2007 Elsevier Ltd. All...

  14. EBT-P project status report

    International Nuclear Information System (INIS)

    1983-11-01

    This Elmo Bumpy Torus Report describes the status of the EBT-P Project in September 1983 after phasedown of the Title II design effort. The report is intended to be a principle source of guidance in the event of a decision to resume work on the project

  15. Determination of left ventricular function parameters and myocardial mass: comparison of MRI and EBT

    International Nuclear Information System (INIS)

    Kivelitz, D.E.; Enzweiler, C.N.H.; Wiese, T.H.; Lembcke, A.; Taupitz, M.; Hamm, B.; Borges, A.; Zytowski, M.

    2000-01-01

    Purpose: Comparative volumetric assessment of the left ventricle by magnetic resonance imaging (MRI) and electron beam tomography (EBT) in patients with ischemic and dilated cardiac disease. Methods: Thirty-two patients underwent cine MRI and EBT in the multislice mode. All studies were triggered to the ECG. Left ventricular ejection fraction (EF), enddiastolic (EDV) and endsystolic volume (ESV), and myocardial mass (MM) were determined by 3D-volumetry by MRI and EBT and results were compared. Results: The correlation between MRI and EBT for EF, EDV, ESV, and MM were r=0.86, r=0.95, r=0.95, and r=0.93, respectively. Conclusions: There is an excellent correlation between MRI and EBT in determining left-ventricular parameters. Both methods are suitable for volumetric assessment of the left ventricie. (orig.) [de

  16. Neutron activation in EBT-P

    International Nuclear Information System (INIS)

    Driemeyer, D.E.

    1983-01-01

    Neutron activation due to photoneutron production in the lead shields proposed to protect the EBT-P superconducting coils from excessive x-ray heating was investigated. The photoneutron flux distribution in various EBT-P structural components was calculated for typical upgrade operating conditions using a standard two-dimensional transport model (TWOTRAN). Activity levels were then evaluated for major structural materials using activation cross sections tabulated in the GAMMON library. Activation dose rates in the device enclosure following several days of 8h/day upgrade (90GHz) operation were found to be approx. 6 mrem/h, decaying to <0.25 mrem/h in approx. 3 days. This requires radiation monitoring of all personnel entering the device enclosure during this time, but should not generally restrict hands on access to the device. There is thus no strong motivation to replace lead with another shield material; however, it may be desirable to borate the enclosure walls in order to reduce the effect which impurities might have on activity levels

  17. Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner

    International Nuclear Information System (INIS)

    McCabe, Bradley P.; Speidel, Michael A.; Pike, Tina L.; Van Lysel, Michael S.

    2011-01-01

    Purpose: In this study, newly formulated XR-RV3 GafChromic film was calibrated with National Institute of Standards and Technology (NIST) traceability for measurement of patient skin dose during fluoroscopically guided interventional procedures. Methods: The film was calibrated free-in-air to air kerma levels between 15 and 1100 cGy using four moderately filtered x-ray beam qualities (60, 80, 100, and 120 kVp). The calibration films were scanned with a commercial flatbed document scanner. Film reflective density-to-air kerma calibration curves were constructed for each beam quality, with both the orange and white sides facing the x-ray source. A method to correct for nonuniformity in scanner response (up to 25% depending on position) was developed to enable dose measurement with large films. The response of XR-RV3 film under patient backscattering conditions was examined using on-phantom film exposures and Monte Carlo simulations. Results: The response of XR-RV3 film to a given air kerma depended on kVp and film orientation. For a 200 cGy air kerma exposure with the orange side of the film facing the source, the film response increased by 20% from 60 to 120 kVp. At 500 cGy, the increase was 12%. When 500 cGy exposures were performed with the white side facing the x-ray source, the film response increased by 4.0% (60 kVp) to 9.9% (120 kVp) compared to the orange-facing orientation. On-phantom film measurements and Monte Carlo simulations show that using a NIST-traceable free-in-air calibration curve to determine air kerma in the presence of backscatter results in an error from 2% up to 8% depending on beam quality. The combined uncertainty in the air kerma measurement from the calibration curves and scanner nonuniformity correction was ±7.1% (95% C.I.). The film showed notable stability. Calibrations of film and scanner separated by 1 yr differed by 1.0%. Conclusions: XR-RV3 radiochromic film response to a given air kerma shows dependence on beam quality and film

  18. LET dependence of GafChromic films and an ion chamber in low-energy proton dosimetry

    International Nuclear Information System (INIS)

    Kirby, Daniel; Parker, David; Green, Stuart; Hugtenburg, Richard; Wojnecki, Cecile; Palmans, Hugo

    2010-01-01

    Dosimetry using a PMMA phantom was performed in 15 and 29 MeV proton beams from the Birmingham cyclotron, with a Markus parallel-plate ionization chamber and GafChromic EBT and MD-V2-55 film. Simulations of the depth-dose curves were performed with FLUKA 2008.3 and MCNPX 2.5.0, which agreed almost perfectly with each other in range and only differed by 2% in the Bragg peak (BP) region. FLUKA was also used to calculate k Q factors for Markus chamber measurements as an improvement to the IAEA TRS-398 values in low-energy beams. FLUKA depth-dose simulations overestimate the BP height measured by ion chamber by about 10%, where the initial proton energy spread was estimated by fitting to the slope of the measured BP distal edge. Both GafChromic films showed an under-response in the BP compared to ion chamber; however, EBT exhibits this effect at lower energies than MD-V2-55. A possible reason for this is attributed to the shape and arrangement of the monomer particles being different in the active components of EBT and MD-V2-55. Relative effectiveness (RE) of both films is presented as functions of residual range R res in water and peak proton energy determined by FLUKA, with considerations for the spatial separation of the two active layers in each film. The proton energies at which RE reduces to 90% of maximum film response are 6.7 and 3.2 MeV for MD-V2-55 and EBT, respectively. Additionally, a beam quality correction factor (g Q,Q 0 ) is suggested for both GafChromic films, involving water-to-film stopping power ratios evaluated using ICRU recommendations, and a polymer yield factor G Q 0 /G Q . RE in this work is equated to the reciprocal of the polymer yield factor. The calculated values of (s w,film ) Q /(s w,film ) Q 0 are constant within 2.1% and 1.2% across the proton energy range of 1-300 MeV for EBT and MD-V2-55, respectively, so it is concluded that the polymer yield factor is the dominant factor causing the LET quenching effect.

  19. Properties of Wide-dose-range GafChromic Films for Synchrotron Radiation Facility

    International Nuclear Information System (INIS)

    Nariyama, Nobuteru

    2007-01-01

    GafChromic films have been used at SPring-8 to detect the intensively irradiated parts and protect them from damage by being covered with shield or moved. To extend the usable dose range more widely, a new type of sensitive film EBT was investigated for the introduction. Calibration curves were obtained irradiated with 60Co γ rays and compared with those of other GafChromic films. For the application, these films were set in the white x-ray hutch and the dose distribution was measured. Ratio of doses given by EBT and XT-R indicated the degree of the photon spectrum hardness, which depended on the positions. As a result, dose range from 50 mGy to 300 kGy became available for dose distribution measurements, and a set of films having different energy responses was found to give information of photon spectra

  20. Characterization and review of dosimetric procedures for movies Gafchromic ETB3

    International Nuclear Information System (INIS)

    Perez Azorin, J. F.; Ramos Garcia, L. I.; Ortiz de Zarate Vivanco, R.; Espinosa Sola, J. M.; Celeiro Gonzalez, J.; Lupiani Castellanos, J.; Alcibar-Arechuluaga, C.; Trueba Carayo, I.

    2013-01-01

    The aim of this study is to characterize the new radiochromic EBT3 movies, presentation of dosimetry procedure followed in our department as well as comparison of our the most relevant procedure found in the literature. (Author)

  1. A quantitative method to the analysis of MLC leaf position and speed based on EPID and EBT3 film for dynamic IMRT treatment with different types of MLC.

    Science.gov (United States)

    Li, Yinghui; Chen, Lixin; Zhu, Jinhan; Wang, Bin; Liu, Xiaowei

    2017-07-01

    A quantitative method based on the electronic portal imaging system (EPID) and film was developed for MLC position and speed testing; this method was used for three MLC types (Millennium, MLCi, and Agility MLC). To determine the leaf position, a picket fence designed by the dynamic (DMLC) model was used. The full-width half-maximum (FWHM) values of each gap measured by EPID and EBT3 were converted to the gap width using the FWHM versus nominal gap width relationship. The algorithm developed for the picket fence analysis was able to quantify the gap width, the distance between gaps, and each individual leaf position. To determine the leaf speed, a 0.5 × 20 cm 2 MLC-defined sliding gap was applied across a 14 × 20 cm 2 symmetry field. The linacs ran at a fixed-dose rate. The use of different monitor units (MUs) for this test led to different leaf speeds. The effect of leaf transmission was considered in a speed accuracy analysis. The difference between the EPID and film results for the MLC position is less than 0.1 mm. For the three MLC types, twice the standard deviation (2 SD) is provided; 0.2, 0.4, and 0.4 mm for gap widths of three MLC types, and 0.1, 0.2, and 0.2 mm for distances between gaps. The individual leaf positions deviate from the preset positions within 0.1 mm. The variations in the speed profiles for the EPID and EBT3 results are consistent, but the EPID results are slightly better than the film results. Different speeds were measured for each MLC type. For all three MLC types, speed errors increase with increasing speed. The analysis speeds deviate from the preset speeds within approximately 0.01 cm s -1 . This quantitative analysis of MLC position and speed provides an intuitive evaluation for MLC quality assurance (QA). © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  2. SU-E-T-447: Electronic Brachytherapy (EBT) Treatment of Cervical Cancer - First Clinical Experience

    Energy Technology Data Exchange (ETDEWEB)

    Johnson, D; Johnson, M; Thompson, J; Ahmad, S [University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma (United States); Chan, L; Hausen, H [Xoft Inc., San Jose, CA (United States)

    2014-06-01

    Purpose: To study the first trial patient in which an electronic brachytherapy (EBT) x-ray source is utilized for treatment of cervical cancer. Methods: During patient treatment, a miniaturized x-ray source was used in combination with a customized titanium tandem and ovoid applicator set. The semi-specialized source was modeled with formalisms outlined by AAMP Task Group 43. Multiple models were used to compensate for variable attenuation conditions as a function of source positions. Varian Brachyvision treatment planning software was utilized on CT data sets for dose calculations prior to treatment delivery. The dose was prescribed to “point A” as defined by American Brachytherapy society. Additional treatments plans were created from those clinically utilized in patient care and were recalculated for an existing Ir-192 source model. Dose volume histograms (DVH) and point dose calculations were compared between the modalities for the clinical condition present in patients treated with EBT. Results: Clinical treatment times, though longer than those typically experienced by Ir-192 users, were manageable. Instantaneous dose rates at personal positions within the treatment vault were lower than those measured during intra operative radiation therapy and breast EBT treatments. Due to lower average photon energy in EBT, dose gradients within the treatment plans were as expected steeper than those observed in Ir-192 based brachytherapy. DVH comparisons between Ir-192 and EBT treatments showed an expected decrease in the integral dose to normal tissues of interest for EBT. In comparing plans created for EBT delivery with those calculated for Ir-192, average dose values for EBT were more than 4%, 11%, and 9% lower at predefined bladder, rectum and “point B” positions, respectively. Conclusion: For the first time, we have demonstrated that the utilizing electronic brachytherapy system for tandem and ovoid based treatment of cancer of the cervix is feasible, and

  3. Vacuum system problems of EBT: a steady-state fusion experiment

    International Nuclear Information System (INIS)

    Livesey, R.L.

    1981-01-01

    Many of the vacuum problems faced by EBT will soon be shared by other plasma devices as high-power microwave systems and long pulse lengths become more common. The solutions used on EBT (such as the raised lip with elastomer seal) are not unique; however, experience has shown that microwave-compatible designs must be carefully thought out. All details of the vacuum must be carefully thought out. All details of the vacuum must be carefully screened in advance to insure that microwaves do not leak into pumps or diagnostics where they can cause major damage. Sputter coating, which even now is noticeably present in most pulsed plasma systems, becomes much worse as systems approach steady state. And finally, radiation degradation of components which is presently a minor problem will become significant on high-power microwave-fed devices, such as EBT-P

  4. Skin Dosimetry in Breast Teletherapy on a Phantom Anthropomorphic and Anthropometric Phantom

    Energy Technology Data Exchange (ETDEWEB)

    Batista Nogueira, Luciana [Anatomy and Imaging Department, Federal University of Minas Gerais, Belo Horizonte (Brazil); Lemos Silva, Hugo Leonardo [Santa Casa Hospital, Belo Horizonte (Brazil); Donato da Silva, Sabrina; Passos Ribeiro Campos, Tarcisio [Nuclear Engineering Department, Federal University of Minas Gerais, Belo Horizonte (Brazil)

    2015-07-01

    This paper addresses the breast teletherapy dosimetry. The goal is to evaluate and compare absorbed doses in equivalent skin tissue, TE-skin, of an anthropomorphic and anthropometric breast phantom submitted to breast radiotherapy. The methodology involved the reproduction of a set of tomographic images of the phantom; the elaboration of conformational radiotherapy planning in the SOMAVISION and CadPlan (TPS) software; and the synthetic breast irradiation by parallel opposed fields in 3D conformal teletherapy at 6 MV linear accelerator Clinac-2100 C from VARIAN with prescribed dose (PD) of 180 cGy to the target volume (PTV), referent to the glandular tissue. Radiochromic films EBT2 were selected as dosimeters. Two independent calibration processes of films with solid water Gammex 457 plates and water filled box were produced. Curves of optical density (OD) versus absorbed dose were produced. Dosimeters were positioned in the external region of the breast phantom in contact with TE-skin, area of 4.0 cm{sup 2} each. The irradiation process was prepared in duplicate to check the reproducibility of the technique. The radiochromic films were scanned and their response in RGB (Red, Green, Blue) analyzed by the ImageJ software. The optical density was obtained and converted to dose based on the calibration curves. Thus, the spatial dose distribution in the skin was reproduced. The absorbed doses measured on the radiochromic films in TE-skin showed values between upper and lower quadrants at 9 o'clock in the range of 54% of PD, between the upper and lower quadrants 3 o'clock in the range of 72% and 6 o'clock at the lower quadrant in the range of 68 % of PD. The values are ±64% (p <0.05) according to the TPS. It is concluded that the depth dose measured in solid water plates or water box reproduce equivalent dose values for both calibration processes of the radiochromic films. It was observed that the skin received doses ranging from 50% to 78% of the

  5. Temperature and hydration effects on absorbance spectra and radiation sensitivity of a radiochromic medium

    Energy Technology Data Exchange (ETDEWEB)

    Rink, Alexandra; Lewis, David F.; Varma, Sangya; Vitkin, I. Alex; Jaffray, David A. [Princess Margaret Hospital/Ontario Cancer Institute, Department of Medical Biophysics and Radiation Oncology, University of Toronto, Toronto, Ontario M5G 2M9 (Canada); Advanced Materials Group, International Specialty Products, Inc., Wayne, New Jersey 07470 (United States); Princess Margaret Hospital/Ontario Cancer Institute, Department of Medical Biophysics and Radiation Oncology, University of Toronto, Toronto, Ontario M5G 2M9 (Canada)

    2008-10-15

    The effects of temperature on real time changes in optical density ({Delta}OD) of GAFCHROMIC EBT film were investigated. The spectral peak of maximum change in absorbance ({lambda}{sub max}) was shown to downshift linearly when the temperature of the film was increased from 22 to 38 degree sign C. The {Delta}OD values were also shown to decrease linearly with temperature, and this decrease could not be attributed to the shift in {lambda}{sub max}. A compensation scheme using {lambda}{sub max} and a temperature-dependent correction factor was investigated, but provided limited improvement. Part of the reason may be the fluctuations in hydration of the active component, which were found to affect both position of absorbance peaks and the sensitivity of the film. To test the effect of hydration, laminated and unlaminated films were desiccated. This shifted both the major and minor absorbance peaks in the opposite direction to the change observed with temperature. The desiccated film also exhibited reduced sensitivity to ionizing radiation. Rehydration of the desiccated films did not reverse the effects, but rather gave rise to another form of the polymer with absorbance maxima upshifted further 20 nm. Hence, the spectral characteristics and sensitivity of the film can be dependent on its history, potentially complicating both real-time and conventional radiation dosimetry.

  6. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE/optical-CT 3D dosimetry system

    Energy Technology Data Exchange (ETDEWEB)

    Clift, Corey; Thomas, Andrew; Chang Zheng; Oldham, Mark [Department of Radiation Oncology, Duke University Medical Center, Durham, NC 27710 (United States); Adamovics, John [Department of Chemistry, Rider University, Lawrenceville, NJ 08648 (United States); Das, Indra [Department of Radiation Oncology, Indiana University School of Medicine, Indianapolis, IN 46202 (United States)], E-mail: cclift@montefiore.org

    2010-03-07

    Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (S{sub c,p}), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT film was used for independent verification. Measurements of S{sub c,p} made with PRESAGE and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2-3.6% for PRESAGE, and 1.6-3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE. The advantages of the PRESAGE (registered) system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence.

  7. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE/optical-CT 3D dosimetry system

    International Nuclear Information System (INIS)

    Clift, Corey; Thomas, Andrew; Chang Zheng; Oldham, Mark; Adamovics, John; Das, Indra

    2010-01-01

    Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (S c,p ), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT film was used for independent verification. Measurements of S c,p made with PRESAGE and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2-3.6% for PRESAGE, and 1.6-3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE. The advantages of the PRESAGE (registered) system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence.

  8. LET dependence of GafChromic films and an ion chamber in low-energy proton dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Kirby, Daniel; Parker, David [School of Physics and Astronomy, University of Birmingham, Birmingham, B15 2TT (United Kingdom); Green, Stuart; Hugtenburg, Richard; Wojnecki, Cecile [Department of Medical Physics, University Hospital Birmingham NHS Trust, Birmingham, B15 2TH (United Kingdom); Palmans, Hugo [National Physical Laboratory, Acoustics and Ionizing Radiation, Teddington (United Kingdom)], E-mail: djk191@bham.ac.uk

    2010-01-21

    Dosimetry using a PMMA phantom was performed in 15 and 29 MeV proton beams from the Birmingham cyclotron, with a Markus parallel-plate ionization chamber and GafChromic EBT and MD-V2-55 film. Simulations of the depth-dose curves were performed with FLUKA 2008.3 and MCNPX 2.5.0, which agreed almost perfectly with each other in range and only differed by 2% in the Bragg peak (BP) region. FLUKA was also used to calculate k{sub Q} factors for Markus chamber measurements as an improvement to the IAEA TRS-398 values in low-energy beams. FLUKA depth-dose simulations overestimate the BP height measured by ion chamber by about 10%, where the initial proton energy spread was estimated by fitting to the slope of the measured BP distal edge. Both GafChromic films showed an under-response in the BP compared to ion chamber; however, EBT exhibits this effect at lower energies than MD-V2-55. A possible reason for this is attributed to the shape and arrangement of the monomer particles being different in the active components of EBT and MD-V2-55. Relative effectiveness (RE) of both films is presented as functions of residual range R{sub res} in water and peak proton energy determined by FLUKA, with considerations for the spatial separation of the two active layers in each film. The proton energies at which RE reduces to 90% of maximum film response are 6.7 and 3.2 MeV for MD-V2-55 and EBT, respectively. Additionally, a beam quality correction factor (g{sub Q,Q{sub 0}}) is suggested for both GafChromic films, involving water-to-film stopping power ratios evaluated using ICRU recommendations, and a polymer yield factor G{sub Q{sub 0}}/G{sub Q}. RE in this work is equated to the reciprocal of the polymer yield factor. The calculated values of (s{sub w,film}){sub Q} /(s{sub w,film}){sub Q{sub 0}} are constant within 2.1% and 1.2% across the proton energy range of 1-300 MeV for EBT and MD-V2-55, respectively, so it is concluded that the polymer yield factor is the dominant factor

  9. Photon beam dose distributions for patients with implanted temporary tissue expanders

    Science.gov (United States)

    Asena, A.; Kairn, T.; Crowe, S. B.; Trapp, J. V.

    2015-01-01

    This study examines the effects of temporary tissue expanders (TTEs) on the dose distributions of photon beams in breast cancer radiotherapy treatments. EBT2 radiochromic film and ion chamber measurements were taken to quantify the attenuation and backscatter effects of the inhomogeneity. Results illustrate that the internal magnetic port present in a tissue expander causes a dose reduction of approximately 25% in photon tangent fields immediately downstream of the implant. It was also shown that the silicone elastomer shell of the tissue expander reduced the dose to the target volume by as much as 8%. This work demonstrates the importance for an accurately modelled high-density implant in the treatment planning system for post-mastectomy breast cancer patients.

  10. Direct reading fast microwave interferometer for EBT

    International Nuclear Information System (INIS)

    Uckan, T.

    1984-10-01

    A simple and inexpensive 4-mm direct reading fast (rise time approx. 100 μs) microwave interferometer is described. The system is particularly useful for density measurements on the ELMO Bumpy Torus (EBT) during pulsed operation

  11. Surface dosimetry for breast radiotherapy in the presence of immobilization cast material

    International Nuclear Information System (INIS)

    Kelly, Andrew; Hardcastle, Nicholas; Metcalfe, Peter; Cutajar, Dean; Quinn, Alexandra; Cardoso, Michael; Rosenfeld, Anatoly; Foo, Kerwyn; Barlin, Sheree

    2011-01-01

    Curative breast radiotherapy typically leaves patients with varying degrees of cosmetic damage. One problem interfering with cosmetically acceptable breast radiotherapy is the external contour for large pendulous breasts which often results in high doses to skin folds. Thermoplastic casts are often employed to secure the breasts to maintain setup reproducibility and limit the presence of skin folds. This paper aims to determine changes in surface dose that can be attributed to the use of thermoplastic immobilization casts. Skin dose for a clinical hybrid conformal/IMRT breast plan was measured using radiochromic film and MOSFET detectors at a range of water equivalent depths representative of the different skin layers. The radiochromic film was used as an integrating dosimeter, while the MOSFETs were used for real-time dosimetry to isolate the contribution of skin dose from individual IMRT segments. Strips of film were placed at various locations on the breast and the MOSFETs were used to measure skin dose at 16 positions spaced along the film strips for comparison of data. The results showed an increase in skin dose in the presence of the immobilization cast of up to 45.7% and 62.3% of the skin dose without the immobilization cast present as measured with Gafchromic EBT film and MOSFETs, respectively. The increase in skin dose due to the immobilization cast varied with the angle of beam incidence and was greatest when the beam was normally incident on the phantom. The increase in surface dose with the immobilization cast was greater under entrance dose conditions compared to exit dose conditions.

  12. Surface dosimetry for breast radiotherapy in the presence of immobilization cast material

    Science.gov (United States)

    Kelly, Andrew; Hardcastle, Nicholas; Metcalfe, Peter; Cutajar, Dean; Quinn, Alexandra; Foo, Kerwyn; Cardoso, Michael; Barlin, Sheree; Rosenfeld, Anatoly

    2011-02-01

    Curative breast radiotherapy typically leaves patients with varying degrees of cosmetic damage. One problem interfering with cosmetically acceptable breast radiotherapy is the external contour for large pendulous breasts which often results in high doses to skin folds. Thermoplastic casts are often employed to secure the breasts to maintain setup reproducibility and limit the presence of skin folds. This paper aims to determine changes in surface dose that can be attributed to the use of thermoplastic immobilization casts. Skin dose for a clinical hybrid conformal/IMRT breast plan was measured using radiochromic film and MOSFET detectors at a range of water equivalent depths representative of the different skin layers. The radiochromic film was used as an integrating dosimeter, while the MOSFETs were used for real-time dosimetry to isolate the contribution of skin dose from individual IMRT segments. Strips of film were placed at various locations on the breast and the MOSFETs were used to measure skin dose at 16 positions spaced along the film strips for comparison of data. The results showed an increase in skin dose in the presence of the immobilization cast of up to 45.7% and 62.3% of the skin dose without the immobilization cast present as measured with Gafchromic EBT film and MOSFETs, respectively. The increase in skin dose due to the immobilization cast varied with the angle of beam incidence and was greatest when the beam was normally incident on the phantom. The increase in surface dose with the immobilization cast was greater under entrance dose conditions compared to exit dose conditions.

  13. Electron energy confinement in ELMO Bumpy Torus (EBT)

    International Nuclear Information System (INIS)

    Hiroe, S.; Haste, G.R.; Dandl, R.A.

    1979-06-01

    Using a calibrated, solid-state, soft x-ray detector, the electron temperature and density have been measured over a wide range of operating conditions of ELMO Bumpy Torus (EBT). The empirical relations of the temperature or the density to the microwave power and the ambient pressure have been determined. The toroidally stored energy has been observed to increase as the stored energy of the hot electron annulus increases. The energy confinement time has been obtained for various plasma parameters and has been found to agree with the neoclassical theory. The advantages of EBT collisionless scaling for fusion plasma confinement have been noted, i.e., n/sub e/tau/sub E/ increases as T/sub e/ 1 5 in the collisionless regime

  14. Experimental dosimetric evaluation in pelvis phantom, subjected to prostate radiation therapy protocol at 15 MV Linac

    Energy Technology Data Exchange (ETDEWEB)

    Matos, Andrea Silva Dias de; Campos, Tarcisio P.R., E-mail: campos@nuclear.ufmg.b [Universidade Federal de Minas Gerais (DEN/UFMG), Belo Horizonte, MG (Brazil). Dept. de Engenharia Nuclear; Dias, Humberto Galvao [Centro de Radioterapia Hospital Luxemburgo, Belo Horizonte, MG (Brazil)

    2011-07-01

    Among the existing malignant neoplasia, the prostate cancer is most common among male population. Due to its high incidence and morbidity, there is a need for investment in advanced technology for better treatment associated with research and social mobilization to prevent the disease. As an efficient method of treatment for such tumor, radiation teletherapy brings favorable results for the patient, particularly when the cancer is diagnosed early. There are, however, the needs to assess the absorbed doses that reach the prostate in the radiation protocols in order to certify the treatment efficacy. The present research goal is to obtain the profile of absorbed dose distributed in a synthetic prostate on male pelvis phantom following a standard radiation therapy protocol. The methodology makes use of a NRI made phantom and a 15MV Linac accelerator. This phantom has anthropomorphic and anthropometric features containing the major internal organs, including bone, prostate, intestine, and bladder. The exposition was made in a 15 MV linear accelerator taken the isocenter in four fields as a 'BOX' of opposing beams. The dosimetry was prepared using GafChromic EBT type 2 radiochromic film and calibration in a solid water phantom. The radiochromic films were digitized on the Microtek Scan Maker 6900XL model scanner operating in the transmission mode and optical density readings based on RGB mode in the computer program Imagedig. The absorbance readings were performed in the spectrophotometer SP-220 mark BIOSPECTRO obtaining calibration curves generated by the collected data. The results reproduce the dose distribution generated in two orthogonal radiochromic films positioned onto the synthetic prostate. Discussions regarding the characteristics of the phantom and methods of irradiation in relation to the achieved dose profile will be addressed. (author)

  15. Experimental dosimetric evaluation in pelvis phantom, subjected to prostate radiation therapy protocol at 15 MV Linac

    International Nuclear Information System (INIS)

    Matos, Andrea Silva Dias de; Campos, Tarcisio P.R.

    2011-01-01

    Among the existing malignant neoplasia, the prostate cancer is most common among male population. Due to its high incidence and morbidity, there is a need for investment in advanced technology for better treatment associated with research and social mobilization to prevent the disease. As an efficient method of treatment for such tumor, radiation teletherapy brings favorable results for the patient, particularly when the cancer is diagnosed early. There are, however, the needs to assess the absorbed doses that reach the prostate in the radiation protocols in order to certify the treatment efficacy. The present research goal is to obtain the profile of absorbed dose distributed in a synthetic prostate on male pelvis phantom following a standard radiation therapy protocol. The methodology makes use of a NRI made phantom and a 15MV Linac accelerator. This phantom has anthropomorphic and anthropometric features containing the major internal organs, including bone, prostate, intestine, and bladder. The exposition was made in a 15 MV linear accelerator taken the isocenter in four fields as a 'BOX' of opposing beams. The dosimetry was prepared using GafChromic EBT type 2 radiochromic film and calibration in a solid water phantom. The radiochromic films were digitized on the Microtek Scan Maker 6900XL model scanner operating in the transmission mode and optical density readings based on RGB mode in the computer program Imagedig. The absorbance readings were performed in the spectrophotometer SP-220 mark BIOSPECTRO obtaining calibration curves generated by the collected data. The results reproduce the dose distribution generated in two orthogonal radiochromic films positioned onto the synthetic prostate. Discussions regarding the characteristics of the phantom and methods of irradiation in relation to the achieved dose profile will be addressed. (author)

  16. Radiochromic liquid solution

    International Nuclear Information System (INIS)

    Noakes, J.E.; Culp, R.A.

    1983-01-01

    A radiochromic solution which is sensitive to small dosages of ionizing and ultraviolet radiation is described. It consists of a solution of a leucocyanide dye in a clear polar solvent with enough organic acid added to make the solution at least slightly acidic and responds to radiation by permanently changing color. Up to one half of the solution by weight can be replaced by a second solution of an aromatic solvent and an organic fluor. Another modification of the invention is a solution of a leucocyanide dye in a clear polar solvent having an aromatic group, an organic fluor, and enough organic acid to make the solution at least slightly acidic. (author)

  17. Increasing clinicians' EBT exploration and preparation behavior in youth mental health services by changing organizational culture with ARC.

    Science.gov (United States)

    Glisson, Charles; Williams, Nathaniel J; Hemmelgarn, Anthony; Proctor, Enola; Green, Philip

    2016-01-01

    Clinician EBT exploration and preparation behavior is essential to the ongoing implementation of new EBTs. This study examined the effect of the ARC organizational intervention on clinician EBT exploration and preparation behavior and assessed the mediating role of organizational culture as a linking mechanism. Fourteen community mental health agencies that serve youth in a major Midwestern metropolis along with 475 clinicians who worked in those agencies, were randomly assigned to either the three-year ARC intervention or control condition. Organizational culture was assessed with the Organizational Social Context (OSC) measure at baseline and follow-up. EBT exploration and preparation behavior was measured as clinician participation in nine separate community EBT workshops held over a three-year period. There was 69 percent greater odds (OR = 1.69, p organizational culture mediated the positive effect of the ARC intervention on clinicians' workshop attendance (a × b = .21; 95% CI:LL = .05, UL = .40). Organizational interventions that create proficient mental health agency cultures can increase clinician EBT exploration and preparation behavior that is essential to the ongoing implementation of new EBTs in community youth mental health settings. Copyright © 2015 Elsevier Ltd. All rights reserved.

  18. mEBT: multiple-matching Evidence-based Translator of Murine Genomic Responses for Human Immunity Studies.

    Science.gov (United States)

    Tae, Donghyun; Seok, Junhee

    2018-05-29

    In this paper, we introduce multiple-matching Evidence-based Translator (mEBT) to discover genomic responses from murine expression data for human immune studies, which are significant in the given condition of mice and likely have similar responses in the corresponding condition of human. mEBT is evaluated over multiple data sets and shows improved inter-species agreement. mEBT is expected to be useful for research groups who use murine models to study human immunity. http://cdal.korea.ac.kr/mebt/. jseok14@korea.ac.kr. Supplementary data are available at Bioinformatics online.

  19. Dose distribution perturbation due to a Co-Cr-Mo prosthesis

    International Nuclear Information System (INIS)

    Castro Novais, J.; Rodriguez Rodriguez, C.; Cabello Murillo, E.; Fernandez Leton, P.; Perez Moreno, J. M.; Lopez Fernandez, A.; Ferrando Sanchez, A.; Martinez Gomez, L. C.

    2009-01-01

    Knowledge of the attenuation and interface effects when irradiating metallic prosthesis is necessary for the radiotherapy treatment of patients with this kind of implants. This report studies the dose distribution of a 6 MV photon beam in the vicinity of a 1,5 cm diameter Co-Cr-Mo prosthesis. Measurements with Gafchromic EBT radiochromic films have been made. Two blocks of cut films have been placed next to the prosthesis, one in each side. Forty two films reaching a height of 1 cm have been piled up in each block. A spatial resolution equal to the thickness of one film (0,24 mm) is achieved. The results show 28% attenuation and the production of a 42% overdose at the entrance interface, 12% and 3% at 1 mm and 2 mm away from the prosthesis respectively. A 5 mm build-up region is originated in the exit interface, where the under dose is less than 10%. The knowledge of the transmission factor and the interface effects allows us to assess the dose calculated by the treatment planning system. (Author) 11 refs.

  20. Dose distribution perturbation due to a Co-Cr-Mo prosthesis; Perturbacion introducida en la distribucion de dosis por una protesis de Co-Cr-Mo

    Energy Technology Data Exchange (ETDEWEB)

    Castro Novais, J.; Rodriguez Rodriguez, C.; Cabello Murillo, E.; Fernandez Leton, P.; Perez Moreno, J. M.; Lopez Fernandez, A.; Ferrando Sanchez, A.; Martinez Gomez, L. C.

    2009-07-01

    Knowledge of the attenuation and interface effects when irradiating metallic prosthesis is necessary for the radiotherapy treatment of patients with this kind of implants. This report studies the dose distribution of a 6 MV photon beam in the vicinity of a 1,5 cm diameter Co-Cr-Mo prosthesis. Measurements with Gafchromic EBT radiochromic films have been made. Two blocks of cut films have been placed next to the prosthesis, one in each side. Forty two films reaching a height of 1 cm have been piled up in each block. A spatial resolution equal to the thickness of one film (0,24 mm) is achieved. The results show 28% attenuation and the production of a 42% overdose at the entrance interface, 12% and 3% at 1 mm and 2 mm away from the prosthesis respectively. A 5 mm build-up region is originated in the exit interface, where the under dose is less than 10%. The knowledge of the transmission factor and the interface effects allows us to assess the dose calculated by the treatment planning system. (Author) 11 refs.

  1. Radiochromic leuco dye micelle hydrogels: I. Initial investigation

    International Nuclear Information System (INIS)

    Jordan, Kevin; Avvakumov, Nikita

    2009-01-01

    This investigation reports the use of surfactants and colorless leuco triarylmethane dyes to form a new class of radiochromic micelle hydrogels for three-dimensional (3D) water-equivalent dosimetry. Gelatin gel samples with several surfactants and leuco dyes were prepared and evaluated for optical transparency, dose sensitivity and diffusion rates. The addition of Triton X-100, a non-ionic surfactant, at levels exceeding the critical micelle concentration provides a transparent hydrogel in which the water insoluble leuco Malachite Green (LMG) can dissolve. During irradiation, the LMG dye precursor converts to Malachite Green (MG + ). The most sensitive reported LMG gel formulation contains 0.3 mM LMG leuco dye, 16 mM trichloroacetic acid, 7 mM Triton X-100 and 4% w/w gelatin. A diffusion coefficient of 0.14 mm 2 h -1 was determined for MG + in this gel by fitting the time-dependent degradation of the transmission profile after irradiating half of the sample. The diffusion rate was three times lower than the standard radiochromic ferrous xylenol-orange (FX) gel. The primary feature of this 3D hydrogel is that it introduces transparent, radiochromic, micelle hydrogels. The radiochromic response to dose is instantaneous and images are stable for several hours. A dosimetric characterization revealed that the dose response is reproducible to within 10% over five separate batches and independent of both energy and dose rate. Uniform pre-irradiation of samples to 5 Gy provided a subsequent near linear response to greater than 110 Gy. LMG gels when read with a fast optical CT scanner can provide full 3D dose distributions in less than 30 min post-irradiation. LMG micelle gels scanned with a 633 nm light source are a promising system for quantitative water- or tissue-equivalent 3D dose verification in the 5-100 Gy dose range. These gels are useful for the scanning of larger volume dosimeters (i.e. >15 cm diameter) since they are easily prepared with inexpensive ingredients

  2. EBT data acquisition and analysis system

    International Nuclear Information System (INIS)

    Burris, R.D.; Greenwood, D.E.; Stanton, J.S.; Geoffroy, K.A.

    1980-10-01

    This document describes the design and implementation of a data acquisition and analysis system for the EBT fusion experiment. The system includes data acquisition on five computers, automatic transmission of that data to a large, central data base, and a powerful data retrieval system. The system is flexible and easy to use, and it provides a fully documented record of the experiments

  3. Hard x-ray measurements of the hot-electron rings in EBT-S

    International Nuclear Information System (INIS)

    Hillis, D.L.

    1982-06-01

    A thorough understanding of the hot electron rings in ELMO Bumpy Torus-Scale (EBT-S) is essential to the bumpy torus concept of plasma production, since the rings provide bulk plasma stability. The hot electrons are produced via electron cyclotron resonant heating using a 28-GHz cw gyrotron, which has operated up to power levels of 200 kW. The parameters of the energetic electron rings are studied via hard x-ray measurement techniques and with diamagnetic pickup coils. The hard x-ray measurements have used collimated NaI(Tl) detectors to determine the electron temperature T/sub e/ and electron density n/sub e/ for the hot electron annulus. Typical values of T/sub e/ are 400 to 500 keV and of n/sub e/ 2 to 5 x 10 11 cm -3 . The total stored energy of a single energetic electron ring as measured by diamagnetic pickup loops approaches approx. 40 J and is in good agreement with that deduced from hard x-ray measurements. By combining the experimental measurements from hard x-rays and the diamagnetic loops, an estimate can be obtained for the volume of a single hot electron ring. The ring volume is determined to be approx. 2.2 litres, and this volume remains approximately constant over the T-mode operating regime. Finally, the power in the electrons scattered out of the ring is measured indirectly by measuring the x-ray radiation produced when those electrons strike the chamber walls. The variation of this radiation with increasing microwave power levels is found to be consistent with classical scattering estimates

  4. Evaluation and mitigation of potential errors in radiochromic film dosimetry due to film curvature at scanning.

    Science.gov (United States)

    Palmer, Antony L; Bradley, David A; Nisbet, Andrew

    2015-03-08

    This work considers a previously overlooked uncertainty present in film dosimetry which results from moderate curvature of films during the scanning process. Small film samples are particularly susceptible to film curling which may be undetected or deemed insignificant. In this study, we consider test cases with controlled induced curvature of film and with film raised horizontally above the scanner plate. We also evaluate the difference in scans of a film irradiated with a typical brachytherapy dose distribution with the film naturally curved and with the film held flat on the scanner. Typical naturally occurring curvature of film at scanning, giving rise to a maximum height 1 to 2 mm above the scan plane, may introduce dose errors of 1% to 4%, and considerably reduce gamma evaluation passing rates when comparing film-measured doses with treatment planning system-calculated dose distributions, a common application of film dosimetry in radiotherapy. The use of a triple-channel dosimetry algorithm appeared to mitigate the error due to film curvature compared to conventional single-channel film dosimetry. The change in pixel value and calibrated reported dose with film curling or height above the scanner plate may be due to variations in illumination characteristics, optical disturbances, or a Callier-type effect. There is a clear requirement for physically flat films at scanning to avoid the introduction of a substantial error source in film dosimetry. Particularly for small film samples, a compression glass plate above the film is recommended to ensure flat-film scanning. This effect has been overlooked to date in the literature.

  5. Multi-institutional evaluation of end-to-end protocol for IMRT/VMAT treatment chains utilizing conventional linacs.

    Science.gov (United States)

    Loughery, Brian; Knill, Cory; Silverstein, Evan; Zakjevskii, Viatcheslav; Masi, Kathryn; Covington, Elizabeth; Snyder, Karen; Song, Kwang; Snyder, Michael

    2018-03-20

    We conducted a multi-institutional assessment of a recently developed end-to-end monthly quality assurance (QA) protocol for external beam radiation therapy treatment chains. This protocol validates the entire treatment chain against a baseline to detect the presence of complex errors not easily found in standard component-based QA methods. Participating physicists from 3 institutions ran the end-to-end protocol on treatment chains that include Imaging and Radiation Oncology Core (IROC)-credentialed linacs. Results were analyzed in the form of American Association of Physicists in Medicine (AAPM) Task Group (TG)-119 so that they may be referenced by future test participants. Optically stimulated luminescent dosimeter (OSLD), EBT3 radiochromic film, and A1SL ion chamber readings were accumulated across 10 test runs. Confidence limits were calculated to determine where 95% of measurements should fall. From calculated confidence limits, 95% of measurements should be within 5% error for OSLDs, 4% error for ionization chambers, and 4% error for (96% relative gamma pass rate) radiochromic film at 3% agreement/3 mm distance to agreement. Data were separated by institution, model of linac, and treatment protocol (intensity-modulated radiation therapy [IMRT] vs volumetric modulated arc therapy [VMAT]). A total of 97% of OSLDs, 98% of ion chambers, and 93% of films were within the confidence limits; measurements were found outside these limits by a maximum of 4%, consistent despite institutional differences in OSLD reading equipment and radiochromic film calibration techniques. Results from this test may be used by clinics for data comparison. Areas of improvement were identified in the end-to-end protocol that can be implemented in an updated version. The consistency of our data demonstrates the reproducibility and ease-of-use of such tests and suggests a potential role for their use in broad end-to-end QA initiatives. Copyright © 2018 American Association of Medical

  6. Objective coding of content and techniques in workplace-based supervision of an EBT in public mental health.

    Science.gov (United States)

    Dorsey, Shannon; Kerns, Suzanne E U; Lucid, Leah; Pullmann, Michael D; Harrison, Julie P; Berliner, Lucy; Thompson, Kelly; Deblinger, Esther

    2018-01-24

    Workplace-based clinical supervision as an implementation strategy to support evidence-based treatment (EBT) in public mental health has received limited research attention. A commonly provided infrastructure support, it may offer a relatively cost-neutral implementation strategy for organizations. However, research has not objectively examined workplace-based supervision of EBT and specifically how it might differ from EBT supervision provided in efficacy and effectiveness trials. Data come from a descriptive study of supervision in the context of a state-funded EBT implementation effort. Verbal interactions from audio recordings of 438 supervision sessions between 28 supervisors and 70 clinicians from 17 public mental health organizations (in 23 offices) were objectively coded for presence and intensity coverage of 29 supervision strategies (16 content and 13 technique items), duration, and temporal focus. Random effects mixed models estimated proportion of variance in content and techniques attributable to the supervisor and clinician levels. Interrater reliability among coders was excellent. EBT cases averaged 12.4 min of supervision per session. Intensity of coverage for EBT content varied, with some discussed frequently at medium or high intensity (exposure) and others infrequently discussed or discussed only at low intensity (behavior management; assigning/reviewing client homework). Other than fidelity assessment, supervision techniques common in treatment trials (e.g., reviewing actual practice, behavioral rehearsal) were used rarely or primarily at low intensity. In general, EBT content clustered more at the clinician level; different techniques clustered at either the clinician or supervisor level. Workplace-based clinical supervision may be a feasible implementation strategy for supporting EBT implementation, yet it differs from supervision in treatment trials. Time allotted per case is limited, compressing time for EBT coverage. Techniques that

  7. Development of 3D Slicer based film dosimetry analysis

    International Nuclear Information System (INIS)

    Alexander, K M; Schreiner, L J; Robinson, A; Pinter, C; Fichtinger, G

    2017-01-01

    Radiochromic film dosimetry has been widely adopted in the clinic as it is a convenient option for dose measurement and verification. Film dosimetry analysis is typically performed using expensive commercial software, or custom made scripts in Matlab. However, common clinical film analysis software is not transparent regarding what corrections/optimizations are running behind the scenes. In this work, an extension to the open-source medical imaging platform 3D Slicer was developed and implemented in our centre for film dosimetry analysis. This extension streamlines importing treatment planning system dose and film imaging data, film calibration, registration, and comparison of 2D dose distributions, enabling greater accessibility to film analysis and higher reliability. (paper)

  8. A verification methodology for in vivo dosimetry in stereotactic radiotherapy; Uma metodologia para verificacao dosimetrica in vivo em radioterapia estereotaxica

    Energy Technology Data Exchange (ETDEWEB)

    Amaral, Leonardo L.; Oliveira, Harley F.; Fairbanks, Leandro R., E-mail: leonardo.fis@usp.br [Universidade de Sao Paulo (HCFMRP/USP), Ribeirao Preto, SP (Brazil). Faculdade de Medicina. Hospital das Clinicas; Nicolucci, Patricia; Netto, Thomaz G. [Universidade de Sao Paulo (FFCLRP/USP), Ribeirao Preto, SP (Brazil). Faculdade de Filosofia, Ciencias e Letras. Departamento de Fisica

    2012-12-15

    Radiotherapy of brain lesions near critical structures requires a high accuracy in the location and dose. The high precision is achieved by the location of the stereotactic apparatus. The accuracy in dose delivery should be accompanied by an accurate quality control in devices that involve the practice, however, still does not guarantee the dose at the time of therapy. The large number of fields and the small size of these conventional methods difficult dosimetry during treatment. The objective of this work was to develop a verification methodology in vivo dosimetry in stereotactic radiotherapy with the aid of the film radiochromic Linear Accelerator with multi leaf collimators Moduleaf. The technique uses film segments radiochromic Gafchromic EBT2, with dimensions of 1x1 cm{sup 2} in area outside the coupled micro-multileaf Moduleaf Siemens. These films were inserted in the region of the central axis of the beam. The films were irradiated and calibrated to obtain the factors that determine the size dependence of the dosimetric field. With these data, we designed a computer program which calculates the density of a film must acquire when subjected to an exposure in this setting. This study evaluated five non-coplanar plans, the first with 15 fields and the other with 25 fields. Before starting the procedure, the film segment is coupled to the device, and after the treatment, the relative density is evaluated and compared with the calculated. The average value of the verification at the time of radiation dosimetry compared with the calculated by the sheet was 1.5%. The data collected in this study showed a satisfactory agreement between measured and calculated by the program in the densitometer. Thus, a methodology was developed to verify in vivo dosimetry in radiotherapy and stereotactic linear accelerator collimators Moduleaf. (author)

  9. SU-G-TeP1-13: Reclined Total Skin Electron Treatment Technique

    Energy Technology Data Exchange (ETDEWEB)

    Mathew, D; Gerbi, B [University of Minnesota, Minneapolis, MN (United States)

    2016-06-15

    Purpose: The purpose is to describe a new reclined technique for treatment of weakened patients that require total skin electron irradiation. Methods: This technique is a modification of a previously published reclined technique differing in that all six patient positions are treated with the gantry angled 60° from vertically down. The patient is located at a treatment distance of 330 cm SSD along the CA of the beam. The 3/8′ thick Lexan beam spoiler is placed 25 cm from the most proximal surface of the patient for all patient treatment positions. To produce a flat, uniform field of ∼190 cm length, the patient was moved longitudinally by an experimentally determined distance. Kodak EDR2 and EBT3 Radiochromic film were placed around the periphery of the phantom, and OSLs were placed every 30° around the phantom periphery to determine output and surface dose uniformity. A piece of Kodak EDR2 was sandwiched between the two slabs of the 30 cm diameter phantom to determine beam penetration. Results: Field uniformity shifting the patient ±75 cm was ±5% over a treatment span of 190 cm. The dose variation around the periphery of the 30 cm diameter phantom varied by <±5% with the maximum values observed at the 0°-300°, 60° locations with the minimum values at the 30°-330°, 60° locations. Results obtained using Kodak EDR2, EBT3 Radiochromic film, and OSLs agreed to within ±5%. Conclusion: This technique provides a very efficient and convenient means by which to treat the entire skin surface of patients incapable of standing for treatment. It provides a treatment field that is both large and uniform enough for adults along with a convenient way to treat four of the six patient treatment positions. The beam spoiler lies to the side of the patient allowing easy access for patient positioning.

  10. SU-E-T-243: MonteCarlo Simulation Study of Polymer and Radiochromic Gel for Three-Dimensional Proton Dose Distribution

    International Nuclear Information System (INIS)

    Park, M; Jung, H; Kim, G; Ji, Y; Kim, K; Park, S

    2014-01-01

    Purpose: To estimate the three dimensional dose distributions in a polymer gel and a radiochromic gel by comparing with the virtual water phantom exposed to proton beams by applying Monte Carlo simulation. Methods: The polymer gel dosimeter is the compositeness material of gelatin, methacrylic acid, hydroquinone, tetrakis, and distilled water. The radiochromic gel is PRESAGE product. The densities of polymer and radiochromic gel were 1.040 and 1.0005 g/cm3, respectively. The shape of water phantom was a hexahedron with the size of 13 × 13 × 15 cm3. The proton beam energies of 72 and 116 MeV were used in the simulation. Proton beam was directed to the top of the phantom with Z-axis and the shape of beam was quadrangle with 10 × 10 cm2 dimension. The Percent depth dose and the dose distribution were evaluated for estimating the dose distribution of proton particle in two gel dosimeters, and compared with the virtual water phantom. Results: The Bragg-peak for proton particles in two gel dosimeters was similar to the virtual water phantom. Bragg-peak regions of polymer gel, radiochromic gel, and virtual water phantom were represented in the identical region (4.3 cm) for 72 MeV proton beam. For 116 MeV proton beam, the Bragg-peak regions of polymer gel, radiochromic gel, and virtual water phantom were represented in 9.9, 9.9 and 9.7 cm, respectively. The dose distribution of proton particles in polymer gel, radiochromic gel, and virtual water phantom was approximately identical in the case of 72 and 116 MeV energies. The errors for the simulation were under 10%. Conclusion: This work indicates the evaluation of three dimensional dose distributions by exposing proton particles to polymer and radiochromic gel dosimeter by comparing with the water phantom. The polymer gel and the radiochromic gel dosimeter show similar dose distributions for the proton beams

  11. Failure analysis of PB-1 (EBTS Be/Cu mockup)

    International Nuclear Information System (INIS)

    Odegard, B.C. Jr.; Cadden, C.H.

    1996-11-01

    Failure analysis was done on PB-1 (series of Be tiles joined to Cu alloy) following a tile failure during a high heat flux experiment in EBTS (electron beam test system). This heat flux load simulated ambient conditions inside ITER; the Be tiles were bonded to the Cu alloy using low-temperature diffusion bonding, which is being considered for fabricating plasma facing components in ITER. Results showed differences between the EBTS failure and a failure during a room temperature tensile test. The latter occurred at the Cu-Be interface in an intermetallic phase formed by reaction of the two metals at the bonding temperature. Fracture strengths measured by these tests were over 300 MPa. The high heat flux specimens failed at the Cu-Cu diffusion bond. Fracture morphology in both cases was a mixed mode of dimple rupture and transgranular cleavage. Several explanations for this difference in failure mechanism are suggested

  12. Establishment of postal audit system in intensity-modulated radiotherapy by radiophotoluminescent glass dosimeters and a radiochromic film.

    Science.gov (United States)

    Okamoto, Hiroyuki; Minemura, Toshiyuki; Nakamura, Mitsuhiro; Mizuno, Hideyuki; Tohyama, Naoki; Nishio, Teiji; Wakita, Akihisa; Nakamura, Satoshi; Nishioka, Shie; Iijima, Kotaro; Fujiyama, Daisuke; Itami, Jun; Nishimura, Yasumasa

    2018-04-01

    We developed an efficient postal audit system to independently assess the delivered dose using radiophotoluminescent glass dosimeters (RPLDs) and the positional differences of fields using EBT3 film at the axial plane for intensity-modulated radiotherapy (IMRT). The audit phantom had a C-shaped target structure as a planning target volume (PTV) with four measurement points for the RPLDs and a cylindrical structure as the organ at risk (OAR) for one measurement point. The phantoms were sent to 24 institutions. Point dose measurements with a 0.6 cm 3 PTW farmer chamber were also performed to justify glass dosimetry in IMRT. The measured dose with the RPLDs was compared to the calculated dose in the institution's treatment planning system (TPS). The mean ± 1.96σ of the ratio of the measured dose with the RPLDs to the farmer chamber was 0.997 ± 0.024 with no significant difference (p = .175). The investigations demonstrated that glass dosimetry was reliable with a high measurement accuracy comparable to the chamber. The mean ± 1.96σ for the dose differences with a reference of the TPS dose for the PTV and the OAR was 0.1 ± 2.5% and -2.1 ± 17.8%, respectively. The mean ± 1.96σ for the right-left and the anterior-posterior direction was -0.9 ± 2.8 and 0.5 ± 1.4 mm, respectively. This study is the first report to justify glass dosimetry for implementation in IMRT audit in Japan. We demonstrate that our postal audit system has high accuracy with a high-level criterion of 3%/3 mm. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  13. Investigating On Colour Stability Conditions Of Postirradiation Radiochromic Film Dosimeter

    International Nuclear Information System (INIS)

    Nguyen Nguyet Dieu; Doan Binh; Pham Thu Hong; Cao Van Chung; Nguyen Thanh Duoc

    2011-01-01

    B3 dosimeter is a thin film with average thickness of 0.0194 mm, which is supplied by the Gex company, the United States. This dosimeter was influenced by many factors: light, temperature, humidity during and after irradiation process. In fact, B3 film dosimeters will be stable under certain conditions such as tightly sealed packs, controlled irradiation and stored temperature after irradiated. Therefore, investigation of the stability effect of postirradiated B3 film dosimeters on the heating temperature, heating time and storing time is carried out before the absorbed dose is read and followed standard reading procedures. When exposed to ionizing radiation, the dosimeters change from colorless to colour. The absorbed doses are read on a Genesys 20 spectrophotometer at a wavelength of 544 nm. Absorbed dose range is investigated from 0.55 to 80 kGy. Experimental results were indicated that colour stability of the postirradiated dosimeters at a temperature of 65 ± 3 o C for 30 minutes and keeping them in desiccator for 5 minutes before read out. Under these conditions, colour stability of B3 film dosimeter has maintained for 3 months. (author)

  14. S3C: EBT Steady-State Shooting code description and user's guide

    International Nuclear Information System (INIS)

    Downum, W.B.

    1983-09-01

    The Oak Ridge National Laboratory (ORNL) one-dimensional (1-D) Steady-State Shooting code (S3C) for ELMO Bumpy Torus (EBT) plasmas is described. Benchmark calculations finding the steady-state density and electron and ion temperature profiles for a known neutral density profile and known external energy sources are carried out. Good agreement is obtained with results from the ORNL Radially Resolved Time Dependent 1-D Transport code for an EBT-Q type reactor. The program logic is described, along with the physics models in each code block and the variable names used. Sample input and output files are listed, along with the main code

  15. Microwave transport in EBT distribution manifolds using Monte Carlo ray-tracing techniques

    International Nuclear Information System (INIS)

    Lillie, R.A.; White, T.L.; Gabriel, T.A.; Alsmiller, R.G. Jr.

    1983-01-01

    Ray tracing Monte Carlo calculations have been carried out using an existing Monte Carlo radiation transport code to obtain estimates of the microsave power exiting the torus coupling links in EPT microwave manifolds. The microwave power loss and polarization at surface reflections were accounted for by treating the microwaves as plane waves reflecting off plane surfaces. Agreement on the order of 10% was obtained between the measured and calculated output power distribution for an existing EBT-S toroidal manifold. A cost effective iterative procedure utilizing the Monte Carlo history data was implemented to predict design changes which could produce increased manifold efficiency and improved output power uniformity

  16. A new method for dosimetry with films radiochromic

    International Nuclear Information System (INIS)

    Mendez Carot, I.

    2013-01-01

    in this paper a new method is presented and the results of the comparison between the calibration is summarized based on a planning reference and calibration obtained from the irradiated fragments measure different dose levels multichannel compare dosimetry based on the weighted average dosimetry described by Micke et al.(present in the FilmQAPro software) and, finally, show different results obtained with the method proposed in several applications clinics. (Author)

  17. Stability of EBT of guiding-centre fluid theory

    International Nuclear Information System (INIS)

    Miller, R.L.

    1981-01-01

    The stability of the hot-electron annulus in the ELMO Bumpy Torus (EBT) is not yet completely understood despite considerable attention. Most stability studies have dealt with localized analysis of simplified models in which the actual magnetic configuration is replaced by a straight-line slab with a gravity to emulate the effects of curvature and gradients in the actual magnetic field. Here, a more realistic geometry, a 'bumpy' cylinder with a 2:1 magnetic mirror ratio, is considered and the response of the hot-electron rings to various non-local perturbations, specifying only the mode number in the ignorable co-ordinate, is examined. Guiding-centre theory (with psub(perpendicular) > psub(parallel)) is used and the second variation in the plasma energy (σW) using a finite-element representation to identify the least stable mode for the plasma is studied. All the equilibria that are examined are found to be unstable for all poloidal mode numbers m>=1, with growth rates increasing with increasing ring beta, plasma beta, and poloidal mode number. It is concluded that two-fluid and/or kinetic effects are required to explain the observed global stability of EBT-I. (author)

  18. Comparison of calculational methods for EBT reactor nucleonics

    International Nuclear Information System (INIS)

    Henninger, R.J.; Seed, T.J.; Soran, P.D.; Dudziak, D.J.

    1980-01-01

    Nucleonic calculations for a preliminary conceptual design of the first wall/blanket/shield/coil assembly for an EBT reactor are described. Two-dimensional Monte Carlo, and one- and two-dimensional discrete-ordinates calculations are compared. Good agreement for the calculated values of tritium breeding and nuclear heating is seen. We find that the three methods are all useful and complementary as a design of this type evolves

  19. Farmers' Market Manager's Level of Communication and Influence on Electronic Benefits Transfer (EBT) Adoption at Midwest Farmers' Markets.

    Science.gov (United States)

    Hasin, Afroza; Smith, Sylvia

    2018-01-01

    To understand market managers' level of communication and use of technology that might influence decision to adopt Electronic Benefits Transfer (EBT) at farmers' markets. Cross-sectional study using the Theory of Diffusion of Innovation. Electronic survey administered in midwest states of Illinois, Michigan, and Wisconsin. Farmers' market managers in Illinois, Michigan, and Wisconsin. Information on EBT adoption, market managers' communication, and technology use. Binary logistic regression analysis with EBT adoption as the dependent variable and frequency of technology use, partnership with organizations, farmers' market association (FMA) membership, Facebook page and Web site for the market, and primary source of information as independent variables. Chi-square tests and ANOVA were used to compare states and adopter categories. Logistic regression results showed that the odds of adopting EBT was 7.5 times higher for markets that had partnership with other organizations. Compared with non-adopters, a significantly greater number of early adopters had partnership, FMA membership, and a Facebook page and Web site for market, and reported to a board of directors. Markets that had partnership, FMA membership, a Facebook page and Web site, and mandatory reporting to a board of directors were important factors that influenced EBT adoption at midwest farmers' markets. Copyright © 2017 Society for Nutrition Education and Behavior. Published by Elsevier Inc. All rights reserved.

  20. Comparative measurement of the neutral density and particle confinement time in EBT

    International Nuclear Information System (INIS)

    Glowienka, J.C.; Richards, R.K.

    1985-11-01

    The neutral density and particle confinement time in the ELMO Bumpy Torus-Scale Experiment (EBT-S) have been determined by two different techniques. These involve a spectroscopic measurement of molecular and atomic hydrogen emissions and a time-decay measurement of a fast-ion population using a diagnostic neutral beam. The results from both diagnostics exhibit identical trends for either estimate, although the absolute values differ by a factor of 2 to 3. The observed variations with fill gas pressure and microwave power from either technique are consistent with measurements of electron density and temperature. In this paper, the measurement techniques are discussed, and the results are compared in the context of consistency with independently observed plasma behavior. 6 refs., 7 figs

  1. SU-E-T-541: Bolus Effect of Thermoplastic Masks in IMRT and VMAT Head and Neck Treatments

    International Nuclear Information System (INIS)

    Zhen, H; Nedzi, L; Chen, S; Jiang, S; Zhao, B

    2014-01-01

    Purpose: To quantitatively evaluate the bolus effect of thermoplalstic mask on patient skin dose during multi-field IMRT and VMAT treatment. Methods: The clinically approved target contours for five head and neck patients were deformably registered to an anthropomorphic Rando phantom. Two plans: Multifield IMRT plan with 7-9 beams and VMAT plan with 2-4 arcs were created for each patient following same dose constraints. 3mm skin was excluded from PTVs but not constrained during optimization. The prescription dose was 200-220 cGy/fraction. A thermoplastic head and shoulder mask was customized for the Rando phantom. Each plan was delivered to the phantom twice with and without mask. During each delivery, two rectangular strips of EBT3 films (1cm x 6.8cm) were placed across the anterior upper and lower neck near PTVs to measure the surface dose. For consistency films were positioned at same locations for same patient. A total of 8 film strips were obtained for each patient. Film dose was calibrated in the range of 0-400cGy on the day of plan delivery. For dose comparison 3 regions of interests (ROIs) of 1×1 cm 2 were selected at left, right and middle part of each film, resulting in 6 point doses at each plan delivery. Results: The films without mask show relatively uniform dose distribution while those with mask clearly show mesh pattern of mask, usually indicating an increase in skin dose. On average the increase in skin dose over all ROIs with mask was 31.9%(±14.8%) with a range of 11.4%- 58.4%. There is no statistically significant difference (p=0.44) between skin dose increase in VMAT (30.8%±15.3%) and IMRT delivery (33.0%±14.9%). Conclusion: Thermoplastic immobilization masks increase surface dose for HN patient by around 30%. The magnitude is comparable between multi-field IMRT and VMAT. Radiochromic EBT3 film serves as an effective tool to quantify bolus effect

  2. Radiation therapy with laser-driven accelerated particle beams: physical dosimetry and spatial dose distribution

    Energy Technology Data Exchange (ETDEWEB)

    Reinhardt, Sabine; Assmann, Walter [Ludwig-Maximilians Universitaet Muenchen (Germany); Kneschaurek, Peter; Wilkens, Jan [MRI, Technische Universitaet Muenchen (Germany)

    2011-07-01

    One of the main goals of the Munich Centre for Advanced Photonics (MAP) is the application of laser driven accelerated (LDA) particle beams for radiation therapy. Due to the unique acceleration process ultrashort particle pulses of high intensity (> 10{sup 7} particles /cm{sup 2}/ns) are generated, which makes online detection an ambitious task. So far, state of the art detection of laser accelerated ion pulses are non-electronic detectors like radiochromic films (RCF), imaging plates (IP) or nuclear track detectors (e.g. CR39). All these kind of detectors are offline detectors requiring several hours of processing time. For this reason they are not qualified for an application in radiation therapy where quantitative real time detection of the beam is an essential prerequisite. Therefore we are investigating pixel detectors for real time monitoring of LDA particle pulses. First tests of commercially available systems with 8-20 MeV protons are presented. For radiobiological experiments second generation Gafchromic films (EBT2) have been calibrated with protons of 12 and 20 MeV for a dose range of 0.3-10 Gy. Dose verification in proton irradiation of subcutaneous tumours in mice was successfully accomplished using these films.

  3. Development of a radiochromic ferric oligomer hydrogel

    International Nuclear Information System (INIS)

    Jordan, Kevin; Sekimoto, Masaya

    2010-01-01

    Ferrous gelatin hydrogels were prepared by using sulphuric acid concentrations lower than required to maintain radiation induced ferric ions fully hydrated. The ferric hydroxyl species that are produced following irradiation exhibit a radiochromic response that can be probed with blue light. The dose distribution shapes were stable in time, indicating no long term diffusion. An over response to dose gradients was observed both in one centimeter cuvette samples and litre volumes probed with optical cone beam CT. This ferrous hydrogel may represent a model system for studying iron radiochemistry in biological systems.

  4. Experimental determination of the radial dose distribution in high gradient regions around 192Ir wires: Comparison of electron paramagnetic resonance imaging, films, and Monte Carlo simulations

    International Nuclear Information System (INIS)

    Kolbun, N.; Leveque, Ph.; Abboud, F.; Bol, A.; Vynckier, S.; Gallez, B.

    2010-01-01

    Purpose: The experimental determination of doses at proximal distances from radioactive sources is difficult because of the steepness of the dose gradient. The goal of this study was to determine the relative radial dose distribution for a low dose rate 192 Ir wire source using electron paramagnetic resonance imaging (EPRI) and to compare the results to those obtained using Gafchromic EBT film dosimetry and Monte Carlo (MC) simulations. Methods: Lithium formate and ammonium formate were chosen as the EPR dosimetric materials and were used to form cylindrical phantoms. The dose distribution of the stable radiation-induced free radicals in the lithium formate and ammonium formate phantoms was assessed by EPRI. EBT films were also inserted inside in ammonium formate phantoms for comparison. MC simulation was performed using the MCNP4C2 software code. Results: The radical signal in irradiated ammonium formate is contained in a single narrow EPR line, with an EPR peak-to-peak linewidth narrower than that of lithium formate (∼0.64 and 1.4 mT, respectively). The spatial resolution of EPR images was enhanced by a factor of 2.3 using ammonium formate compared to lithium formate because its linewidth is about 0.75 mT narrower than that of lithium formate. The EPRI results were consistent to within 1% with those of Gafchromic EBT films and MC simulations at distances from 1.0 to 2.9 mm. The radial dose values obtained by EPRI were about 4% lower at distances from 2.9 to 4.0 mm than those determined by MC simulation and EBT film dosimetry. Conclusions: Ammonium formate is a suitable material under certain conditions for use in brachytherapy dosimetry using EPRI. In this study, the authors demonstrated that the EPRI technique allows the estimation of the relative radial dose distribution at short distances for a 192 Ir wire source.

  5. Bottoms up design of the Elmo Bumpy Torus - proof of principal (EBT-P) fusion research facility

    International Nuclear Information System (INIS)

    Erickson, D.T.

    1981-01-01

    The McDonnell Douglas Astronautics Company, under subcontract to the Union Carbide Corporation Nuclear Division at the DOE Oak Ridge National Laboratory has committed to furnish the EBT-P research facility. Gilbert Associates, Inc. of Reading, Pennysylvania, as the Architect and Engineering subcontractor has been selected for design and construction of this facility. The bottoms up effort to provide the EBT-P facility is now alive and well, with the property purchased, dedication ceremonies conducted, the Preliminary Design effort completed and detail design currently active

  6. Software design for the EBT-P data acquisition and control system R and D

    International Nuclear Information System (INIS)

    Boyd, R.A.

    1983-01-01

    The instrumentation and control system for the EBT-P device is composed of a hierarchy of programmable logic controllers, microprocessor-based data acquisition computers, and a large minicomputer-based facility computer system. The software being developed to support this data acquisition and control system is necessarily quite complex due to several requirements imposed upon the EBT-P overall design criteria. These requirements, which include such considerations as overall reliability, operator interface, real-time display, interprocessor communication, and minimum cost to build, operate, and maintain, dictate that the software be developed in a well structured and controlled manner. To this end, structured software engineering practices are being applied to the design and development of the EBT-P data acquistion and control software. The design process began with the production of a software Requirements Document which describes the hardware and software environment in which the software development takes place. It identifies the major deliverable software items to be produced and describes the practices to be used to design and develop the software. The software design is split into three components: the facility computer software, the microcomputer software, and the PLC software. Within these physical boundaries, the following five functions are defined: data acquisition, display, communication, storage, and control. The software design is further detailed in a Structured Specification Document for each of the three physical components. Each specification describes the software in detailed terms so that a programmer can directly write the required software. Each specification is composed of: data flow diagrams, a data dictionary, structure diagrams, and program design language mini-specifications. Examples of the design issues exposed and addressed during the structured decomposition of EBT-P software processes are discussed in detail

  7. Dosimetric verification of IMRT plans

    International Nuclear Information System (INIS)

    Bulski, W.; Cheimicski, K.; Rostkowska, J.

    2012-01-01

    Intensity modulated radiotherapy (IMRT) is a complex procedure requiring proper dosimetric verification. IMRT dose distributions are characterized by steep dose gradients which enable to spare organs at risk and allow for an escalation of the dose to the tumor. They require large number of radiation beams (sometimes over 10). The fluence measurements for individual beams are not sufficient for evaluation of the total dose distribution and to assure patient safety. The methods used at the Centre of Oncology in Warsaw are presented. In order to measure dose distributions in various cross-sections the film dosimeters were used (radiographic Kodak EDR2 films and radiochromic Gafchromic EBT films). The film characteristics were carefully examined. Several types of tissue equivalent phantoms were developed. A methodology of comparing measured dose distributions against the distributions calculated by treatment planning systems (TPS) was developed and tested. The tolerance level for this comparison was set at 3% difference in dose and 3 mm in distance to agreement. The so called gamma formalism was used. The results of these comparisons for a group of over 600 patients are presented. Agreement was found in 87 % of cases. This film dosimetry methodology was used as a benchmark to test and validate the performance of commercially available 2D and 3D matrices of detectors (ionization chambers or diodes). The results of these validations are also presented. (authors)

  8. Verification of status of reference of Brainscan planning system through Grafchromic EBT3; Verificacion del estado de referencia del sistema de planificacion Brainscan mediante pelicula radiocromica Grafchromic EBT3

    Energy Technology Data Exchange (ETDEWEB)

    Vazquez Vazquez, R.; Ambroa Rey, E. M.; Gimenez Insua, M.; Luna Vega, V.; Mosquera Sueiro, J.; Lobato Busto, R.; Sanchez Garcia, M.; Otero Martinez, C.; Pombar Camean, M.

    2013-07-01

    To compare the State of current of the Brainscan planner with the initial reference State has been successful. This new detector shows behavior similar to its predecessor (Gafchromic EBT2), but without the disadvantages arising from the construction of the same. (Author)

  9. The incidence of the different sources of noise on the uncertainty in radiochromic film dosimetry using single channel and multichannel methods

    Science.gov (United States)

    González-López, Antonio; Vera-Sánchez, Juan Antonio; Ruiz-Morales, Carmen

    2017-11-01

    The influence of the various sources of noise on the uncertainty in radiochromic film (RCF) dosimetry using single channel and multichannel methods is investigated in this work. These sources of noise are extracted from pixel value (PV) readings and dose maps. Pieces of an RCF were each irradiated to different uniform doses, ranging from 0 to 1092 cGy. Then, the pieces were read at two resolutions (72 and 150 ppp) with two flatbed scanners: Epson 10000XL and Epson V800, representing two states of technology. Noise was extracted as described in ISO 15739 (2013), separating its distinct constituents: random noise and fixed pattern (FP) noise. Regarding the PV maps, FP noise is the main source of noise for both models of digitizer. Also, the standard deviation of the random noise in the 10000XL model is almost twice that of the V800 model. In the dose maps, the FP noise is smaller in the multichannel method than in the single channel ones. However, random noise is higher in this method, throughout the dose range. In the multichannel method, FP noise is reduced, as a consequence of this method’s ability to eliminate channel independent perturbations. However, the random noise increases, because the dose is calculated as a linear combination of the doses obtained by the single channel methods. The values of the coefficients of this linear combination are obtained in the present study, and the root of the sum of their squares is shown to range between 0.9 and 1.9 over the dose range studied. These results indicate the random noise to play a fundamental role in the uncertainty of RCF dosimetry: low levels of random noise are required in the digitizer to fully exploit the advantages of the multichannel dosimetry method. This is particularly important for measuring high doses at high spatial resolutions.

  10. Control/interlock/display system for EBT-P using commercially-available hardware and firmware

    International Nuclear Information System (INIS)

    Schmitt, R.J.

    1983-01-01

    For the EBT-P project, alternative commercially-available hardware, software and firmware have been employed for control, interlock and data display functions. This paper describes the criteria and rationale used to select that commercial equipment and discusses the important features of the equipment chosen, especially programmable controllers. Additional discussion is centered on interface problems which are encountered upon attempts to integrate equipment from several vendors. Some solutions to these problems are discussed. Details of software and hardware performance during tests are presented. The extent to which the EBT-P hardware and software configuration addresses and resolves various issues is discussed. Several areas have been uncovered in which relatively slight improvements/modifications of commercial programmable controller firmware would significantly improve the capability of this type of hardware in fusion control applications. These improvements are discussed in detail

  11. Automatic analysis of intrinsic positional verification films brachytherapy using MATLAB

    International Nuclear Information System (INIS)

    Quiros Higueras, J. D.; Marco Blancas, N. de; Ruiz Rodriguez, J. C.

    2011-01-01

    One of the essential tests in quality control of brachytherapy equipment is verification auto load intrinsic positional radioactive source. A classic method for evaluation is the use of x-ray film and measuring the distance between the marks left by autoradiography of the source with respect to a reference. In our center has developed an automated method of measurement by the radiochromic film scanning and implementation of a macro developed in Matlab, in order to optimize time and reduce uncertainty in the measurement. The purpose of this paper is to describe the method developed, assess their uncertainty and quantify their advantages over the manual method. (Author)

  12. A practical three-dimensional dosimetry system for radiation therapy

    International Nuclear Information System (INIS)

    Guo Pengyi; Adamovics, John; Oldham, Mark

    2006-01-01

    There is a pressing need for a practical three-dimensional (3D) dosimetry system, convenient for clinical use, and with the accuracy and resolution to enable comprehensive verification of the complex dose distributions typical of modern radiation therapy. Here we introduce a dosimetry system that can achieve this challenge, consisting of a radiochromic dosimeter (PRESAGE trade mark sign ) and a commercial optical computed tomography (CT) scanning system (OCTOPUS trade mark sign ). PRESAGE trade mark sign is a transparent material with compelling properties for dosimetry, including insensitivity of the dose response to atmospheric exposure, a solid texture negating the need for an external container (reducing edge effects), and amenability to accurate optical CT scanning due to radiochromic optical contrast as opposed to light-scattering contrast. An evaluation of the performance and viability of the PRESAGE trade mark sign /OCTOPUS, combination for routine clinical 3D dosimetry is presented. The performance of the two components (scanner and dosimeter) was investigated separately prior to full system test. The optical CT scanner has a spatial resolution of ≤1 mm, geometric accuracy within 1 mm, and high reconstruction linearity (with a R 2 value of 0.9979 and a standard error of estimation of ∼1%) relative to independent measurement. The overall performance of the PRESAGE trade mark sign /OCTOPUS system was evaluated with respect to a simple known 3D dose distribution, by comparison with GAFCHROMIC[reg] EBT film and the calculated dose from a commissioned planning system. The 'measured' dose distribution in a cylindrical PRESAGE trade mark sign dosimeter (16 cm diameter and 11 cm height) was determined by optical-CT, using a filtered backprojection reconstruction algorithm. A three-way Gamma map comparison (4% dose difference and 4 mm distance to agreement), between the PRESAGE trade mark sign , EBT and calculated dose distributions, showed full agreement in

  13. Dosimetry auditing procedure with alanine dosimeters for light ion beam therapy

    DEFF Research Database (Denmark)

    Ableitinger, Alexander; Vatnitsky, Stanislav; Herrmann, Rochus

    2013-01-01

    Background and purpose In the next few years the number of facilities providing ion beam therapy with scanning beams will increase. An auditing process based on an end-to-end test (including CT imaging, planning and dose delivery) could help new ion therapy centres to validate their entire logistic...... of the biological dose is out of scope of the current work. Materials and methods The audit procedure was based on a homogeneous phantom that mimics the dimension of a head (20 × 20 × 21 cm3). The phantom can be loaded either with an ionisation chamber or 20 alanine dosimeters plus 2 radiochromic EBT films. Dose...... fluence in the alanine detector. A pilot run was performed with protons and carbon ions at the Heidelberg Ion Therapy facility (HIT). Results The mean difference of the absolute physical dose measured with the alanine dosimeters compared with the expected dose from the treatment planning system was −2...

  14. Verification of status of reference of Brainscan planning system through Grafchromic EBT3

    International Nuclear Information System (INIS)

    Vazquez Vazquez, R.; Ambroa Rey, E. M.; Gimenez Insua, M.; Luna Vega, V.; Mosquera Sueiro, J.; Lobato Busto, R.; Sanchez Garcia, M.; Otero Martinez, C.; Pombar Camean, M.

    2013-01-01

    To compare the State of current of the Brainscan planner with the initial reference State has been successful. This new detector shows behavior similar to its predecessor (Gafchromic EBT2), but without the disadvantages arising from the construction of the same. (Author)

  15. Experimental dosimetric evaluation of inhomogeneity effects caused by thoracic vertebrae; Avaliacao dosimetrica experimental de efeitos de inomogeneidade causados por vertebras toracicas

    Energy Technology Data Exchange (ETDEWEB)

    Castro, Andre L.S.; Thompson, Larissa; Campos, Tarcisio P.R., E-mail: radioterapia.andre@gmail.com [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear; Instituto de Radioterapia Sao Francisco, Belo Horizonte, MG (Brazil); Centro de Tratamento em Radioterapia, Betim, MG (Brazil)

    2017-11-01

    The presence of tissue inhomogeneities alters the absorbed dose distribution, which magnitude depends on the physical properties of these tissues and the quality of the radiation. Incorrect assessment of dose distribution may affect local tumor control or increase the normal tissue complication probabilities. The aim of this study was to investigate the influence of the thoracic vertebrae inhomogeneous effects on the dose absorbed by the surrounding soft tissue. The values predicted by the treatment planning system (TPS) were compared to the experimental measurements with EBT-2 radiochromic films positioned on a simulator consisting of only water and inserted axially into a thoracic phantom made of synthetic bone and water. There was a significant change in the dose distribution pattern, increased absorbed dose at the bone-soft tissue interface and high point doses adjacent to the bone compared to the results obtained for the films in homogeneous medium and TPS. The experimental measurements in the water agreed with the TPS within 1.0% with respect to the modal dose whereas the biggest difference found for the medium containing the vertebrae was of 4.6%, however, both values are within the experimental uncertainty. (author)

  16. Preliminary validation of a Monte Carlo model for IMRT fields

    International Nuclear Information System (INIS)

    Wright, Tracy; Lye, Jessica; Mohammadi, Mohammad

    2011-01-01

    Full text: A Monte Carlo model of an Elekta linac, validated for medium to large (10-30 cm) symmetric fields, has been investigated for small, irregular and asymmetric fields suitable for IMRT treatments. The model has been validated with field segments using radiochromic film in solid water. The modelled positions of the multileaf collimator (MLC) leaves have been validated using EBT film, In the model, electrons with a narrow energy spectrum are incident on the target and all components of the linac head are included. The MLC is modelled using the EGSnrc MLCE component module. For the validation, a number of single complex IMRT segments with dimensions approximately 1-8 cm were delivered to film in solid water (see Fig, I), The same segments were modelled using EGSnrc by adjusting the MLC leaf positions in the model validated for 10 cm symmetric fields. Dose distributions along the centre of each MLC leaf as determined by both methods were compared. A picket fence test was also performed to confirm the MLC leaf positions. 95% of the points in the modelled dose distribution along the leaf axis agree with the film measurement to within 1%/1 mm for dose difference and distance to agreement. Areas of most deviation occur in the penumbra region. A system has been developed to calculate the MLC leaf positions in the model for any planned field size.

  17. Dye film dosimetry for radiation processing

    International Nuclear Information System (INIS)

    Humphreys, J.C.; McLaughlin, W.L.

    1981-01-01

    Commercially available plastic films containing dyes or dye precursors are convenient dosimeters and imaging media for electron beams or photons used for industrial radiation processing. As ''grainless'' imaging systems having thicknesses down to a few micrometers, they provide high spatial resolution for determining detailed absorbed dose distributions through microdensitometric analysis. The radiation absorption properties of these systems are adjusted by changing film composition so that the dosimeter materials can be made to simulate the material of interest undergoing irradiation. Other advantages include long-term stability, dose-rate independence, and ease of use and calibration. Radiochromic dye films with thicknesses varying from 0.005 to 1 mm are presently used to monitor electron-beam or gamma-ray doses from 10 to 10 5 Gy (10 3 to 10 7 rad), typical of those encountered in medical applications, radiation curing of polymeric composites, wire and cable insulation, shrinkable plastic tubing and film, as well as sterilization of medical supplies and treatment of municipal and industrial wastes. An NBS calibration service to industry involves the traceability of standard 60 Co gamma ray absorbed dose measurements by means of these films employed as transfer standards

  18. Depth dependence of the single chamber response function of the I'mRT MatriXX array in a 6 MV photon beam

    International Nuclear Information System (INIS)

    Alashrah, Saleh

    2013-01-01

    One of the factors which influence the spatial resolution of a 2D detector array is the size of the single detector, another the transport of the secondary electrons from the walls into the measuring volume. In this study, the single ion chamber dose response function of an I'mRT MatriXX array was determined by comparison between slit beam dose profiles measured with the array and with EBT2 radiochromic film in a solid water-equivalent phantom at a shallow depth of 0.5 cm and at a depth of 5 cm beyond the depth dose maximum for a 6 MV photon beam. The dose response functions were obtained using two methods, the best fit method and the deconvolution method. At the shallow depth, a Lorentz function and at 5 cm depth a Gaussian function, both with the same FWHM of 7.4 mm within limits of uncertainty, were identified as the best suited dose response functions of the 4.5 mm diameter single array chamber. These dose response functions were then tested on various dose profiles whose true shape had been determined with EBT2 film and with the IC03 ionization chamber. By convolving these with the Lorentz kernel (at shallow depth) and the Gaussian kernel (at 5 cm depth) the signal profiles measured with the I'mRT MatriXX array were closely approximated. Thus, the convolution of TPS-calculated dose profiles with these dose response functions can minimize the differences between calculation and measurement which occur due to the limited spatial resolution of the I'mRT MatriXX detector. (orig.)

  19. Patient Dose During Carotid Artery Stenting With Embolic-Protection Devices: Evaluation With Radiochromic Films and Related Diagnostic Reference Levels According to Factors Influencing the Procedure

    International Nuclear Information System (INIS)

    D’Ercole, Loredana; Quaretti, Pietro; Cionfoli, Nicola; Klersy, Catherine; Bocchiola, Milena; Rodolico, Giuseppe; Azzaretti, Andrea; Lisciandro, Francesco; Cascella, Tommaso; Zappoli Thyrion, Federico

    2013-01-01

    To measure the maximum entrance skin dose (MESD) on patients undergoing carotid artery stenting (CAS) using embolic-protection devices, to analyze the dependence of dose and exposure parameters on anatomical, clinical, and technical factors affecting the procedure complexity, to obtain some local diagnostic reference levels (DRLs), and to evaluate whether overcoming DRLs is related to procedure complexity. MESD were evaluated with radiochromic films in 31 patients (mean age 72 ± 7 years). Five of 33 (15 %) procedures used proximal EPD, and 28 of 33 (85 %) procedures used distal EPD. Local DRLs were derived from the recorded exposure parameters in 93 patients (65 men and 28 women, mean age 73 ± 9 years) undergoing 96 CAS with proximal (33 %) or distal (67 %) EPD. Four bilateral lesions were included. MESD values (mean 0.96 ± 0.42 Gy) were FR ) were 269 Gy cm 2 , 28 minutes, and 251, respectively. Only simultaneous bilateral treatment was associated with KAP (odds ratio [OR] 10.14, 95 % confidence interval [CI] 1–102.7, p FR overexposures (OR 10.8, 95 % CI 1.1–109.5, p FR overexposure (OR 2.8, 95 % CI 1.1–7.4, p = 0.040). At multivariable analysis, stenosis ≥ 90 % (OR 2.8, 95 % CI 1.1–7.4, p = 0.040) and bilateral treatment (OR 10.8, 95 % CI 1.1–109.5, p = 0.027) were associated with overexposure for two or more parameters. Skin doses are not problematic in CAS with EPD because these procedures rarely lead to doses >2 Gy.

  20. Radiation-chemical formation of acids in polyvinyl butyral films with chlorinated additives

    International Nuclear Information System (INIS)

    Kriminiskaya, Z.K.

    1993-01-01

    Radiochromic indicators are commonly produced by reacting an indicator dye with an acid formed inside a polymer by irradiation. Halogenated and unhalogenated polymers were used, the latter containing halogenated organics. It was therefore of interest to study the formation of acid in polyvinyl butyral (PVD) with addition of a halogenated compound. Yields were measured of radiation-chemical acid formation in PVB films containing chloral hydrate and hexachloroethane. 5 refs., 1 fig., 2 tabs

  1. Investigation of nanoscale structures by small-angle X-ray scattering in a radiochromic dosimeter

    DEFF Research Database (Denmark)

    Skyt, Peter Sandegaard; Jensen, Grethe Vestergaard; Wahlstedt, Isak Hannes

    2014-01-01

    This study examines the nanoscale structures in a radiochromic dosimeter that was based on leuco-malachite-green dye and the surfactant sodium dodecyl sulfate (SDS) suspended in a gelatin matrix. Small-angle X-ray scattering was used to investigate the structures of a range of compositions...

  2. Development of a head phantom to be used for quality control in stereotactic radiosurgery; Desenvolvimento de um simulador de cabeca para a aplicacao no controle da qualidade de radiocirurgia estereotaxica

    Energy Technology Data Exchange (ETDEWEB)

    Barbosa, Nilseia Aparecida

    2010-05-15

    It was designed and developed a geometric acrylic head phantom (GHP) for Quality Assurance (QA) in Stereotactic Radiosurgery (SRS). Inside the phantom there are inserts that are able to accommodate acrylic targets representing the tumor tissue and organ at risk in the region cranial brain, the brain stem. The tumor tissue is represented by two semi-spheres of acrylic with a diameter of 13.0 mm and cavities in the central region for accommodation of a TLD-100 detector and a small radiochromic EBT Gafchromic filmstrip. The brain stem is represented by the two parts of acrylic cylinder with a diameter 18.0 mm, 38.0 mm length and cavities along the central region to accommodate the 5 detectors TLD-100 and yet another of EBT film. The distance tumor - brain stem is 2.0 mm. The experimental setup was filled with water, attached to the stereotactic frame to determine the coordinates of the target and underwent computed tomography (CT). Cf images were transferred to the SRS planning system BrainLab (BrainScan). The contours of the lesion and organ at risk were delineated and, through the technique of multiple circular arcs, the planning was conduced with five arches, one isocenter and a collimator of 17.5 mm from the combination between the table and gantry . The dose delivered to the isocenter of the lesion was 3.0 Gy and the total coverage of tumor volume corresponds to the 75% isodose. This experimental arrangement is subjected to radiosurgery treatment, after which the dosimeters are evaluated and their responses compared with the values of planned doses. The linear accelerator used was a Varian CLlNAC 2300 CID, photon beam of 6 MV, installed at the National Cancer Institute (INCA). For verification of dose distributions in 3D, the films were irradiated in three planes: sagittal, caronal and axial. The .films were scanned and digitized on a scanner Microtek ScanMaker 9800XL model. The dose distributions in irradiated films were compared with the distributions of doses

  3. Development of a head phantom to be used for quality control in stereotactic radiosurgery

    International Nuclear Information System (INIS)

    Barbosa, Nilseia Aparecida

    2010-05-01

    It was designed and developed a geometric acrylic head phantom (GHP) for Quality Assurance (QA) in Stereotactic Radiosurgery (SRS). Inside the phantom there are inserts that are able to accommodate acrylic targets representing the tumor tissue and organ at risk in the region cranial brain, the brain stem. The tumor tissue is represented by two semi-spheres of acrylic with a diameter of 13.0 mm and cavities in the central region for accommodation of a TLD-100 detector and a small radiochromic EBT Gafchromic filmstrip. The brain stem is represented by the two parts of acrylic cylinder with a diameter 18.0 mm, 38.0 mm length and cavities along the central region to accommodate the 5 detectors TLD-100 and yet another of EBT film. The distance tumor - brain stem is 2.0 mm. The experimental setup was filled with water, attached to the stereotactic frame to determine the coordinates of the target and underwent computed tomography (CT). Cf images were transferred to the SRS planning system BrainLab (BrainScan). The contours of the lesion and organ at risk were delineated and, through the technique of multiple circular arcs, the planning was conduced with five arches, one isocenter and a collimator of 17.5 mm from the combination between the table and gantry . The dose delivered to the isocenter of the lesion was 3.0 Gy and the total coverage of tumor volume corresponds to the 75% isodose. This experimental arrangement is subjected to radiosurgery treatment, after which the dosimeters are evaluated and their responses compared with the values of planned doses. The linear accelerator used was a Varian CLlNAC 2300 CID, photon beam of 6 MV, installed at the National Cancer Institute (INCA). For verification of dose distributions in 3D, the films were irradiated in three planes: sagittal, caronal and axial. The .films were scanned and digitized on a scanner Microtek ScanMaker 9800XL model. The dose distributions in irradiated films were compared with the distributions of doses

  4. EBT time-dependent point model code: description and user's guide

    International Nuclear Information System (INIS)

    Roberts, J.F.; Uckan, N.A.

    1977-07-01

    A D-T time-dependent point model has been developed to assess the energy balance in an EBT reactor plasma. Flexibility is retained in the model to permit more recent data to be incorporated as they become available from the theoretical and experimental studies. This report includes the physics models involved, the program logic, and a description of the variables and routines used. All the files necessary for execution are listed, and the code, including a post-execution plotting routine, is discussed

  5. SU-E-T-753: Three-Dimensional Dose Distributions of Incident Proton Particle in the Polymer Gel Dosimeter and the Radiochromic Gel Dosimeter: A Simulation Study with MCNP Code

    International Nuclear Information System (INIS)

    Park, M; Kim, G; Ji, Y; Kim, K; Park, S; Jung, H

    2015-01-01

    Purpose: The purpose of this study is to estimate the three-dimensional dose distributions in the polymer and the radiochromic gel dosimeter, and to identify the detectability of both gel dosimeters by comparing with the water phantom in case of irradiating the proton particles. Methods: The normoxic polymer gel and the LCV micelle radiochromic gel were used in this study. The densities of polymer and the radiochromic gel dosimeter were 1.024 and 1.005 g/cm 3 , respectively. The dose distributions of protons in the polymer and radiochromic gel were simulated using Monte Carlo radiation transport code (MCNPX, Los Alamos National Laboratory). The shape of phantom irradiated by proton particles was a hexahedron with the dimension of 12.4 × 12.4 × 15.0 cm 3 . The energies of proton beam were 50, 80, and 140 MeV energies were directed to top of the surface of phantom. The cross-sectional view of proton dose distribution in both gel dosimeters was estimated with the water phantom and evaluated by the gamma evaluation method. In addition, the absorbed dose(Gy) was also calculated for evaluating the proton detectability. Results: The evaluation results show that dose distributions in both gel dosimeters at intermediated section and Bragg-peak region are similar with that of the water phantom. At entrance section, however, inconsistencies of dose distribution are represented, compared with water. The relative absorbed doses in radiochromic and polymer gel dosimeter were represented to be 0.47 % and 2.26 % difference, respectively. These results show that the radiochromic gel dosimeter was better matched than the water phantom in the absorbed dose evaluation. Conclusion: The polymer and the radiochromic gel dosimeter show similar characteristics in dose distributions for the proton beams at intermediate section and Bragg-peak region. Moreover the calculated absorbed dose in both gel dosimeters represents similar tendency by comparing with that in water phantom

  6. In-vivo dosimetry with Gafchromic films for multi-isocentric VMAT irradiation of total marrow lymph-nodes: a feasibility study

    International Nuclear Information System (INIS)

    Mancosu, Pietro; Navarria, Pierina; Reggiori, Giacomo; Cozzi, Luca; Fogliata, Antonella; Gaudino, Anna; Lobefalo, Francesca; Paganini, Lucia; Palumbo, Valentina; Sarina, Barbara; Stravato, Antonella; Castagna, Luca; Tomatis, Stefano; Scorsetti, Marta

    2015-01-01

    Total marrow (lymph-nodes) irradiation (TMI-TMLI) by volumetric modulated arc therapy (VMAT) was shown to be feasible by dosimetric feasibility studies. It was demonstrated that several partially overlapping arcs with different isocenters are required to achieve the desired coverage of the hematopoietic or lymphoid tissues targets and to spare the neighbouring healthy tissues. The effect of isocenter shifts was investigated with the treatment planning system but an in- vivo verification of the procedure was not carried out. The objective of this study was the in-vivo verification of the consistency between the delivered and planned doses using bi-dimensional GafChromic EBT3 films. In a first phase a phantom study was carried out to quantify the uncertainties under controlled conditions. In a second phase three patients treated with TMLI were enrolled for in-vivo dosimetry. The dose prescription was 2Gy in single fraction. Ten arcs paired on 4-6 isocenters were used to cover the target. Cone Beam Computed Tomography (CBCT) was used to verify the patient positioning at each isocenter. GafChromic EBT3 films were placed below the patient on the top of a dedicated immobilization system specifically designed. The dose maps measured with the EBT3 films were compared with the corresponding calculations along the patient support couch. Gamma Agreement Index (GAI) with dose difference of 5% and distance to agreement of 5 mm was computed. In the phantom study, optimal target coverage and healthy tissue sparing was observed. GAI(5%,5 mm) was 99.4%. For the patient-specific measurements, GAI(5%,5 mm) was greater than 95% and GAI (5%,3 mm) > 90% for all patients. In vivo measurements demonstrated the delivered dose to be in good agreement with the planned one for the TMI-TMLI protocol where partially overlapping arcs with different isocenters are required

  7. Two-point model for electron transport in EBT

    International Nuclear Information System (INIS)

    Chiu, S.C.; Guest, G.E.

    1980-01-01

    The electron transport in EBT is simulated by a two-point model corresponding to the central plasma and the edge. The central plasma is assumed to obey neoclassical collisionless transport. The edge plasma is assumed turbulent and modeled by Bohm diffusion. The steady-state temperatures and densities in both regions are obtained as functions of neutral influx and microwave power. It is found that as the neutral influx decreases and power increases, the edge density decreases while the core density increases. We conclude that if ring instability is responsible for the T-M mode transition, and if stability is correlated with cold electron density at the edge, it will depend sensitively on ambient gas pressure and microwave power

  8. End-to-end tests using alanine dosimetry in scanned proton beams

    Science.gov (United States)

    Carlino, A.; Gouldstone, C.; Kragl, G.; Traneus, E.; Marrale, M.; Vatnitsky, S.; Stock, M.; Palmans, H.

    2018-03-01

    This paper describes end-to-end test procedures as the last fundamental step of medical commissioning before starting clinical operation of the MedAustron synchrotron-based pencil beam scanning (PBS) therapy facility with protons. One in-house homogeneous phantom and two anthropomorphic heterogeneous (head and pelvis) phantoms were used for end-to-end tests at MedAustron. The phantoms were equipped with alanine detectors, radiochromic films and ionization chambers. The correction for the ‘quenching’ effect of alanine pellets was implemented in the Monte Carlo platform of the evaluation version of RayStation TPS. During the end-to-end tests, the phantoms were transferred through the workflow like real patients to simulate the entire clinical workflow: immobilization, imaging, treatment planning and dose delivery. Different clinical scenarios of increasing complexity were simulated: delivery of a single beam, two oblique beams without and with range shifter. In addition to the dose comparison in the plastic phantoms the dose obtained from alanine pellet readings was compared with the dose determined with the Farmer ionization chamber in water. A consistent systematic deviation of about 2% was found between alanine dosimetry and the ionization chamber dosimetry in water and plastic materials. Acceptable agreement of planned and delivered doses was observed together with consistent and reproducible results of the end-to-end testing performed with different dosimetric techniques (alanine detectors, ionization chambers and EBT3 radiochromic films). The results confirmed the adequate implementation and integration of the new PBS technology at MedAustron. This work demonstrates that alanine pellets are suitable detectors for end-to-end tests in proton beam therapy and the developed procedures with customized anthropomorphic phantoms can be used to support implementation of PBS technology in clinical practice.

  9. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/ optical-CT 3D dosimetry system

    Science.gov (United States)

    Clift, Corey; Thomas, Andrew; Adamovics, John; Chang, Zheng; Das, Indra; Oldham, Mark

    2010-03-01

    Achieving accurate small field dosimetry is challenging. This study investigates the utility of a radiochromic plastic PRESAGE® read with optical-CT for the acquisition of radiosurgery field commissioning data from a Novalis Tx system with a high-definition multileaf collimator (HDMLC). Total scatter factors (Sc, p), beam profiles, and penumbrae were measured for five different radiosurgery fields (5, 10, 20, 30 and 40 mm) using a commercially available optical-CT scanner (OCTOPUS, MGS Research). The percent depth dose (PDD), beam profile and penumbra of the 10 mm field were also measured using a higher resolution in-house prototype CCD-based scanner. Gafchromic EBT® film was used for independent verification. Measurements of Sc, p made with PRESAGE® and film agreed with mini-ion chamber commissioning data to within 4% for every field (range 0.2-3.6% for PRESAGE®, and 1.6-3.6% for EBT). PDD, beam profile and penumbra measurements made with the two PRESAGE®/optical-CT systems and film showed good agreement with the high-resolution diode commissioning measurements with a competitive resolution (0.5 mm pixels). The in-house prototype optical-CT scanner allowed much finer resolution compared with previous applications of PRESAGE®. The advantages of the PRESAGE® system for small field dosimetry include 3D measurements, negligible volume averaging, directional insensitivity, an absence of beam perturbations, energy and dose rate independence.

  10. One and one-half dimensional model of the EBT reactor

    International Nuclear Information System (INIS)

    Klein, H.H.; Bathke, C.G.

    1979-01-01

    A one-dimensional, time-dependent model is described for plasma particle and energy transport and alpha particle transport coupled with magnetic field evolution in a geometry appropriate to EBT. The transport equations used are derived from exact moments of the Boltzmann equation, and the magnetic field is calculated from Faraday's and Ampere's laws. The set of transport equations is closed by incorporating into them transport coefficiencents derived from the appropriate kinetic equation. Also included in the model is a Fokker-Planck calculation of the alpha particle slowing down and resultant plasma heating

  11. Experimental determination of the effective point of measurement for various detectors used in photon and electron beam dosimetry.

    Science.gov (United States)

    Looe, Hui Khee; Harder, Dietrich; Poppe, Björn

    2011-07-21

    The subject of this study is the 'shift of the effective point of measurement', Δz, well known as a method of correction compensating for the 'displacement effect' in photon and electron beam dosimetry. Radiochromic EBT 1 films have been used to measure the 'true' TPR curves of 6 and 15 MV photons and 6 and 9 MeV electrons in the solid water-equivalent material RW3. For the Roos and Markus chambers, the cylindrical 'PinPoint', 'Semiflex' and 'Rigid-Stem' chambers, the 2D-Array and the E-type silicon diode (all from PTW-Freiburg), the positions of the effective points of measurement have been determined by direct or indirect comparison between their TPR curves and those of the EBT 1 film. Both for the Roos and Markus chambers, we found Δz = (0.4 ± 0.1) mm, which confirms earlier experimental and Monte Carlo results, but means a shortcoming of the 'water-equivalent window thickness' formula. For the cylindrical chambers, the ratio Δz/r was observed to increase with r, confirming a recent Monte Carlo prediction by Tessier (2010 E2-CN-182, Paper no 147, IDOS, Vienna) as well as the experimental observations by Johansson et al (1978 IAEA Symp. Proc. (Vienna) IAEA-SM-222/35 pp 243-70). According to a theoretical consideration, the shift of the effective point of measurement from the reference point of the detector is caused by a gradient of the fluence of the ionizing particles. As the experiments have shown, the value of Δz depends on the construction of the detector, but remains invariant under changes of radiation quality and depth. Other disturbances, which do not belong to the class of 'gradient effects', are not corrected by shifting the effective point of measurement.

  12. Clinical evaluation of an endorectal immobilization system for use in prostate hypofractionated Stereotactic Ablative Body Radiotherapy (SABR)

    International Nuclear Information System (INIS)

    Nicolae, Alexandru; Davidson, Melanie; Easton, Harry; Helou, Joelle; Musunuru, Hima; Loblaw, Andrew; Ravi, Ananth

    2015-01-01

    The objective of this study was to evaluate a novel prostate endorectal immobilization system (EIS) for improving the delivery of hypofractionated Stereotactic Ablative Body Radiotherapy (SABR) for prostate cancer. Twenty patients (n = 20) with low- or intermediate-risk prostate cancer (T1-T2b, Gleason Score < 7, PSA ≤ 20 ng/mL), were treated with an EIS in place using Volumetric Modulated Arc Therapy (VMAT), to a prescription dose of 26 Gy delivered in 2 fractions once per week; the intent of the institutional clinical trial was an attempt to replicate brachytherapy-like dosimetry using SABR. EBT3 radiochromic film embedded within the EIS was used as a quality assurance measure of the delivered dose; additionally, prostate intrafraction motion captured using pre- and post-treatment conebeam computed tomography (CBCT) scans was evaluated. Treatment plans were generated for patients with- and without the EIS to evaluate its effects on target and rectal dosimetry. None of the observed 3-dimensional prostate displacements were ≥ 3 mm over the elapsed treatment time. A Gamma passing rate of 95.64 ± 4.28 % was observed between planned and delivered dose profiles on EBT3 film analysis in the low-dose region. No statistically significant differences between treatment plans with- and without-EIS were observed for rectal, bladder, clinical target volume (CTV), and PTV contours (p = 0.477, 0.484, 0.487, and 0.487, respectively). A mean rectal V80% of 1.07 cc was achieved for plans using the EIS. The EIS enables the safe delivery of brachytherapy-like SABR plans to the prostate while having minimal impact on treatment planning and rectal dosimetry. Consistent and reproducible immobilization of the prostate is possible throughout the duration of these treatments using such a device

  13. Fabrication of a flexible polycarbonate/porphyrin film dosimeter for high dose dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Feizi, Shahzad [Nuclear Science and Technology Research Institute (NSTRI), Tehran (Iran, Islamic Republic of). Radiation Application Research School

    2017-10-01

    Dyed polycarbonate (PC) Radiochromic films with 20 μm thickness were prepared by casting of organic solution of PC containing 0.5 wt.% tetrakis (pentafluorophenyl) porphyrin (TPPF{sub 20}) on a glass petri dish. Characterization of the film as a routine dosimeter was studied. On subjecting PC/TPPF{sub 20} film dosimeter to gamma radiation, a gradual decrease in the color of films was observed. The sensitivity of these films and the linearity of dose-response curves were studied under {sup 60}Co γ-rays expose in dose range of 0-100 kGy. The results were compared with the commercial and non-commercial dosimeters. Experimental parameters including humidity, temperature and pre-irradiation (shelf-life) and post-irradiation storage in dark and in indirect sunlight were examined. The maximum absorbance of soret band of TPPF{sub 20} had a bathochromic shift and appeared at 414 nm which remained intact in the investigated dose range. The dyed films characteristics were found to be stable enough in media with high degrees of temperature and humidity. The results indicate that radiation induced decoloration of PC/TPPF{sub 20} films can be reliably used in high dose dosimetry.

  14. Effects of temperature and humidity during irradiation on the response of radiachromic film dosimeters

    International Nuclear Information System (INIS)

    Ningnoi, T.; Ehlermann, D.A.E.

    1994-01-01

    The effects of temperature and humidity during γ irradiation on the response of two types of film dosimeters (Far West radiochromic and GafChromic films) were studied in the dose range of 0.3-3 kGy. Both films show a significant effect of temperature and humidity and a simple correction function is proposed. This correction is usually between 5 and 10% for the range studied. For the GafChromic film, a colour change at temperatures above 50 o C was observed and, consequently, this system cannot be used at these temperatures. At lower temperatures down to -70 o C the sensitivity of both films is reduced and a simple correction is possible. In this study and for the dose ranges used, only a slight dependence on humidity was observed for both films from 0 to 60% r.h. Whereas the GafChromic film at humidities up to 90% r.h. shows only a moderate effect, the Far West film shows a considerable inconsistency for the dose range studied. A simple correction function may be applied for humidity effects, except for the Far West film above 60% r.h. where the effect of humidity is also dose dependent. (author)

  15. In vivo dosimetry with MOSFETs and GAFCHROMIC films during electron IORT for Accelerated Partial Breast Irradiation.

    Science.gov (United States)

    Petoukhova, Anna; Rüssel, Iris; Nijst-Brouwers, Julienne; van Wingerden, Ko; van Egmond, Jaap; Jacobs, Daphne; Marinelli, Andreas; van der Sijp, Joost; Koper, Peter; Struikmans, Henk

    2017-12-01

    The purpose of this study was to compare the delivered dose to the expected intraoperative radiation therapy (IORT) dose with in vivo dosimetry. For IORT using electrons in accelerated partial breast irradiation, this is especially relevant since a high dose is delivered in a single fraction. For 47 of breast cancer patients, in vivo dosimetry was performed with MOSFETs and/or GAFCHROMIC EBT2 films. A total dose of 23.33 Gy at d max was given directly after completing the lumpectomy procedure with electron beams generated with an IORT dedicated mobile accelerator. A protection disk was used to shield the thoracic wall. The results of in vivo MOSFET dosimetry for 27 patients and GAFROMIC film dosimetry for 20 patients were analysed. The entry dose for the breast tissue, measured with MOSFETs, (mean value 22.3 Gy, SD 3.4%) agreed within 1.7% with the expected dose (mean value 21.9 Gy). The dose in breast tissue, measured with GAFCHROMIC films (mean value 23.50 Gy) was on average within 0.7% (SD = 3.7%, range -5.5% to 5.6%) of the prescribed dose of 23.33 Gy. The dose measured with MOSFETs and GAFROMIC EBT2 films agreed well with the expected dose. For both methods, the dose to the thoracic wall, lungs and heart for left sided patents was lower than 2.5 Gy even when 12 MeV was applied. The positioning time of GAFCHROMIC films is negligible and based on our results we recommend its use as a standard tool for patient quality assurance during breast cancer IORT. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  16. SU-E-T-282: Dose Measurements with An End-To-End Audit Phantom for Stereotactic Radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Jones, R; Artschan, R [Calvary Mater Newcastle, Newcastle, NSW (Australia); Thwaites, D [University of Sydney, Sydney, NSW (Australia); Lehmann, J [Calvary Mater Newcastle, Newcastle, NSW (Australia); University of Sydney, Sydney, NSW (Australia)

    2015-06-15

    Purpose: Report on dose measurements as part of an end-to-end test for stereotactic radiotherapy, using a new audit tool, which allows audits to be performed efficiently either by an onsite team or as a postal audit. Methods: Film measurements have been performed with a new Stereotactic Cube Phantom. The phantom has been designed to perform Winston Lutz type position verification measurements and dose measurements in one setup. It comprises a plastic cube with a high density ball in its centre (used for MV imaging with film or EPID) and low density markers in the periphery (used for Cone Beam Computed Tomography, CBCT imaging). It also features strategically placed gold markers near the posterior and right surfaces, which can be used to calculate phantom rotations on MV images. Slit-like openings allow insertion of film or other detectors.The phantom was scanned and small field treatment plans were created. The fields do not traverse any inhomogeneities of the phantom on their paths to the measurement location. The phantom was setup at the delivery system using CBCT imaging. The calculated treatment fields were delivered, each with a piece of radiochromic film (EBT3) placed in the anterior film holder of the phantom. MU had been selected in planning to achieve similar exposures on all films. Calibration films were exposed in solid water for dose levels around the expected doses. Films were scanned and analysed following established procedures. Results: Setup of the cube showed excellent suitability for CBCT 3D alignment. MV imaging with EPID allowed for clear identification of all markers. Film based dose measurements showed good agreement for MLC created fields down to 0.5 mm × 0.5 mm. Conclusion: An end-to-end audit phantom for stereotactic radiotherapy has been developed and tested.

  17. Estimation of eye lens dose during brain scans using Gafchromic XR-QA2 film in various multidetector CT scanners

    International Nuclear Information System (INIS)

    Akhilesh, Philomina; Jamhale, Shramika H.; Sharma, S.D.; Kumar, Rajesh; Datta, D.; Kulkarni, Arti R.

    2017-01-01

    The purpose of this study was to estimate eye lens dose during brain scans in 16-, 64-, 128- and 256-slice multidetector computed tomography (CT) scanners in helical acquisition mode and to test the feasibility of using radiochromic film as eye lens dosemeter during CT scanning. Eye lens dose measurements were performed using Gafchromic XR-QA2 film on a polystyrene head phantom designed with outer dimensions equivalent to the head size of a reference Indian man. The response accuracy of XR-QA2 film was validated by using thermoluminescence dosemeters. The eye lens dose measured using XR-QA2 film on head phantom for plain brain scanning in helical mode ranged from 43.8 to 45.8 mGy. The XR-QA2 film measured dose values were in agreement with TLD measured dose values within a maximum variation of 8.9%. The good correlation between the two data sets confirms the viability of using XR-QA2 film for eye lens dosimetry. (authors)

  18. Iron-based radiochromic systems for UV dosimetry applications

    Science.gov (United States)

    Lee, Hannah J.; Alqathami, Mamdooh; Blencowe, Anton; Ibbott, Geoffrey

    2018-01-01

    Phototherapy treatment using ultraviolet (UV) A and B light sources has long existed as a treatment option for various skin conditions. Quality control for phototherapy treatment recommended by the British Association of Dermatologists and British Photodermatology Group generally focused on instrumentation-based dosimetry measurements. The purpose of this study was to present an alternative, easily prepared dosimeter system for the measurement of UV dose and as a simple quality assurance technique for phototherapy treatments. Five different UVA-sensitive radiochromic dosimeter formulations were investigated and responded with a measurable and visible optical change both in solution and in gel form. Iron(III) reduction reaction formulations were found to be more sensitive to UVA compared to iron(II) oxidation formulations. One iron(III) reduction formulation was found to be especially promising due to its sensitivity to UVA dose, ease of production, and linear response up to a saturation point.

  19. Evaluation of effective energy for QA and QC: measurement of half-value layer using radiochromic film density

    International Nuclear Information System (INIS)

    Gotanda, R.; Takeda, Y.; Gotanda, T.; Oishi Hospital, Hiroshima; Tabuchi, A.; Kawasaki Hospital, Okayama; Yamamoto, K.; Osaka Cancer Prevention and Detection Centre, Osaka; Kuwano, T.; Osaka Medical Center for Cancer and Cardovascular Diseases, Osaka; Yatake, H.; Kaizuka City Hospital, Osaka; Katsuda, T.

    2009-01-01

    The effective energy of diagnostic x-rays is important for quality assurance (QA) and quality control (QC). However, the half-value layer (HVL), which is necessary to evaluate the effective energy, is not ubiquitously monitored because ionization-chamber dosimetry is time-consuming and complicated. To verify the applicability of GAFCHROMIC XR type R (GAF-R) film for HVL measurement as an alternative to monitoring with an ionization chamber, a single-strip method for measuring the HVL has been evaluated. Calibration curves of absorbed dose versus film density were generated using this single-strip method with GAF-R film, and the coefficient of determination (r2) of the straight-line approximation was evaluated. The HVLs (effective energies) estimated using the GAF-R film and an ionization chamber were compared. The coefficient of determination (r2) of the straight-line approximation obtained with the GAF-R film was more than 0.99. The effective energies (HVLs) evaluated using the GAF-R film and the ionization chamber were 43.25 keV (5.10 m m) and 39.86 keV (4.45 mm), respectively. The difference in the effective energies determined by the two methods was thus 8.5%. These results suggest that GAF-R might be used to evaluate the effective energy from the film-density growth without the need for ionization-chamber measurements.

  20. Experimental measurements of spatial dose distributions in radiosurgery treatments

    International Nuclear Information System (INIS)

    Avila-Rodriguez, M. A.; Rodriguez-Villafuerte, M.; Diaz-Perches, R.; Perez-Pastenes, M. A.

    2001-01-01

    The measurement of stereotactic radiosurgery dose distributions requires an integrating, high-resolution dosimeter capable of providing a spatial map of absorbed dose. This paper describes the use of a commercial radiochromic dye film (GafChromic MD-55-2) to measure radiosurgery dose distributions with 6 MV X-rays in a head phantom. The response of the MD-55-2 was evaluated by digitizing and analyzing the films with conventional computer systems. Radiosurgery dose distributions were measured using the radiochromic film in a spherical acrylic phantom of 16 cm diameter undergoing a typical SRS treatment as a patient, and were compared with dose distributions provided by the treatment planning system. The comparison lead to mean radial differences of ±0.6 mm, ±0.9 mm, ±1.3 mm, ±1.9 mm, and ±2.8 mm, for the 80, 60, 50, 40, and 30% isodose curves, respectively. It is concluded that the radiochromic film is a convenient and useful tool for radiosurgery treatment planning validation

  1. Patient Dose During Carotid Artery Stenting With Embolic-Protection Devices: Evaluation With Radiochromic Films and Related Diagnostic Reference Levels According to Factors Influencing the Procedure

    Energy Technology Data Exchange (ETDEWEB)

    D' Ercole, Loredana, E-mail: l.dercole@smatteo.pv.it [Fondazione IRCCS Policlinico San Matteo, Department of Medical Physics (Italy); Quaretti, Pietro; Cionfoli, Nicola [Fondazione IRCCS Policlinico San Matteo, Department of Radiology (Italy); Klersy, Catherine [Fondazione IRCCS Policlinico San Matteo, Biometry and Clinical Epidemiology Service, Research Department, (Italy); Bocchiola, Milena [Fondazione IRCCS Policlinico San Matteo, Department of Medical Physics (Italy); Rodolico, Giuseppe; Azzaretti, Andrea [Fondazione IRCCS Policlinico San Matteo, Department of Radiology (Italy); Lisciandro, Francesco [Fondazione IRCCS Policlinico San Matteo, Department of Medical Physics (Italy); Cascella, Tommaso; Zappoli Thyrion, Federico [Fondazione IRCCS Policlinico San Matteo, Department of Radiology (Italy)

    2013-04-15

    To measure the maximum entrance skin dose (MESD) on patients undergoing carotid artery stenting (CAS) using embolic-protection devices, to analyze the dependence of dose and exposure parameters on anatomical, clinical, and technical factors affecting the procedure complexity, to obtain some local diagnostic reference levels (DRLs), and to evaluate whether overcoming DRLs is related to procedure complexity. MESD were evaluated with radiochromic films in 31 patients (mean age 72 {+-} 7 years). Five of 33 (15 %) procedures used proximal EPD, and 28 of 33 (85 %) procedures used distal EPD. Local DRLs were derived from the recorded exposure parameters in 93 patients (65 men and 28 women, mean age 73 {+-} 9 years) undergoing 96 CAS with proximal (33 %) or distal (67 %) EPD. Four bilateral lesions were included. MESD values (mean 0.96 {+-} 0.42 Gy) were <2 Gy without relevant dependence on procedure complexity. Local DRL values for kerma area product (KAP), fluoroscopy time (FT), and number of frames (N{sub FR}) were 269 Gy cm{sup 2}, 28 minutes, and 251, respectively. Only simultaneous bilateral treatment was associated with KAP (odds ratio [OR] 10.14, 95 % confidence interval [CI] 1-102.7, p < 0.05) and N{sub FR} overexposures (OR 10.8, 95 % CI 1.1-109.5, p < 0.05). Type I aortic arch decreased the risk of FT overexposure (OR 0.4, 95 % CI 0.1-0.9, p = 0.042), and stenosis {>=} 90 % increased the risk of N{sub FR} overexposure (OR 2.8, 95 % CI 1.1-7.4, p = 0.040). At multivariable analysis, stenosis {>=} 90 % (OR 2.8, 95 % CI 1.1-7.4, p = 0.040) and bilateral treatment (OR 10.8, 95 % CI 1.1-109.5, p = 0.027) were associated with overexposure for two or more parameters. Skin doses are not problematic in CAS with EPD because these procedures rarely lead to doses >2 Gy.

  2. Physicochemical studies on some new thin films based on diacetylene materials for possible use in radiation processing dosimetry

    International Nuclear Information System (INIS)

    Bayomi, A.M.M.

    2014-01-01

    The aim of the present thesis was to develop new thin radiochromic films, self-adhesive labels, radiation indicators, and gel dosimeters based on some conjugated diacetylene derivatives, as radiation-sensitive components, for radiation processing dosimetry. The diacetylene derivatives used were synthesized first in the laboratory then they were introduced in various mixtures to prepare the dosimeters and labels. The first part of the present thesis includes the synthesis of conjugated diacetylene monomer of 2,4-hexadiyne-1,6-bis(n-butyl urethane) (HDDBU) and the introduction of this monomer into poly(vinyl butyral) (PVB) organic solutions with different concentrations (chapter 4.1.). The prepared PVB solutions were then coated on flexible polyester sheets to prepare a thin film dosimeter. This film was investigated in the dose range of 3-150 kGy by UV-Vis spectrophotometer in the spectrum range of 200-400 nm for potential use in radiation dose process control. Additionally, the x-ray diffraction (XRD) analysis of the synthesized HDDBU monomer and the PVB film containing HDDBU has been performed. In PVB film, HDDBU monomer polymerizes under gamma radiation inducing change in the UV-Vis absorption spectra of the film. The spectrum of unirradiated films featured three absorption bands at 233 nm, 245 nm, and 259 nm, which are characteristic of the diacetylene chromofore. The intensity of these bands grows with the radiation dose without any shift in their position. In addition, two new absorption bands, namely, 273 nm and 285 nm develop upon irradiation, and their intensities increase proportionally with increasing absorbed dose. According to XRD spectroscopy, HDDBU monomer was in a non-crystalline state when it was mixed into PVB film. The PVB film with appropriate concentrations of HDDBU can be used for dose measurements in the range 3 - 150 kGy. They are suitable for use in monitoring various industrial processes, such as food irradiation, radiation sterilization of

  3. Automatic analysis of intrinsic positional verification films brachytherapy using MATLAB; Analisis automatico de peliculas de verificacion posicional intrinsica en braqueterapia mediante MATLAB

    Energy Technology Data Exchange (ETDEWEB)

    Quiros Higueras, J. D.; Marco Blancas, N. de; Ruiz Rodriguez, J. C.

    2011-07-01

    One of the essential tests in quality control of brachytherapy equipment is verification auto load intrinsic positional radioactive source. A classic method for evaluation is the use of x-ray film and measuring the distance between the marks left by autoradiography of the source with respect to a reference. In our center has developed an automated method of measurement by the radiochromic film scanning and implementation of a macro developed in Matlab, in order to optimize time and reduce uncertainty in the measurement. The purpose of this paper is to describe the method developed, assess their uncertainty and quantify their advantages over the manual method. (Author)

  4. Dosimetric comparison of different treatment modalities for stereotactic radiotherapy.

    Science.gov (United States)

    Hsu, Shih-Ming; Lai, Yuan-Chun; Jeng, Chien-Chung; Tseng, Chia-Ying

    2017-09-16

    The modalities for performing stereotactic radiotherapy (SRT) on the brain include the cone-based linear accelerator (linac), the flattening filter-free (FFF) volumetric modulated arc therapy (VMAT) linac, and tomotherapy. In this study, the cone-based linac, FFF-VMAT linac, and tomotherapy modalities were evaluated by measuring the differences in doses delivered during brain SRT and experimentally assessing the accuracy of the output radiation doses through clinical measurements. We employed a homemade acrylic dosimetry phantom representing the head, within which a thermoluminescent dosimeter (TLD) and radiochromic EBT3 film were installed. Using the conformity/gradient index (CGI) and Paddick methods, the quality of the doses delivered by the various SRT modalities was evaluated. The quality indicators included the uniformity, conformity, and gradient indices. TLDs and EBT3 films were used to experimentally assess the accuracy of the SRT dose output. The dose homogeneity indices of all the treatment modalities were lower than 1.25. The cone-based linac had the best conformity for all tumors, regardless of the tumor location and size, followed by the FFF-VMAT linac; tomography was the worst-performing treatment modality in this regard. The cone-based linac had the best gradient, regardless of the tumor location and size, whereas the FFF-VMAT linac had a better gradient than tomotherapy for a large tumor diameter (28 mm). The TLD and EBT3 measurements of the dose at the center of tumors indicated that the average difference between the measurements and the calculated dose was generally less than 4%. When the 3% 3-mm gamma passing rate metric was used, the average passing rates of all three treatment modalities exceeded 98%. Regarding the dose, the cone-based linac had the best conformity and steepest dose gradient for tumors of different sizes and distances from the brainstem. The results of this study suggest that SRT should be performed using the cone

  5. SU-F-T-574: MLC Based SRS Beam Commissioning - Minimum Target Size Investigation

    Energy Technology Data Exchange (ETDEWEB)

    Zakikhani, R [Florida Cancer Specialists - Largo, Largo, FL (United States); Able, C [Florida Cancer Specialists - New Port Richey, New Port Richey, FL (United States)

    2016-06-15

    Purpose: To implement a MLC accelerator based SRS program using small fields down to 1 cm × 1 cm and to determine the smallest target size safe for clinical treatment. Methods: Computerized beam scanning was performed in water using a diode detector and a linac-head attached transmission ion chamber to characterize the small field dosimetric aspects of a 6 MV photon beam (Trilogy-Varian Medical Systems, Inc.). The output factors, PDD and profiles of field sizes 1, 2, 3, 4, and 10 cm{sup 2} were measured and utilized to create a new treatment planning system (TPS) model (AAA ver 11021). Static MLC SRS treatment plans were created and delivered to a homogeneous phantom (Cube 20, CIRS, Inc.) for a 1.0 cm and 1.5 cm “PTV” target. A 12 field DMLC plan was created for a 2.1 cm target. Radiochromic film (EBT3, Ashland Inc.) was used to measure the planar dose in the axial, coronal and sagittal planes. A micro ion chamber (0.007 cc) was used to measure the dose at isocenter for each treatment delivery. Results: The new TPS model was validated by using a tolerance criteria of 2% dose and 2 mm distance to agreement. For fields ≤ 3 cm{sup 2}, the max PDD, Profile and OF difference was 0.9%, 2%/2mm and 1.4% respectively. The measured radiochromic film planar dose distributions had gamma scores of 95.3% or higher using a 3%/2mm criteria. Ion chamber measurements for all 3 test plans effectively met our goal of delivering the dose accurately to within 5% when compared to the expected dose reported by the TPS (1 cm plan Δ= −5.2%, 1.5 cm plan Δ= −2.0%, 2 cm plan Δ= 1.5%). Conclusion: End to end testing confirmed that MLC defined SRS for target sizes ≥ 1.0 cm can be safely planned and delivered.

  6. Dosimetric validation of planning system Eclipse 10 in partial breast irradiation treatments with IMRT; Validacion dosimetrica del sistema de planeacion Eclipse 10 en tratamientos de irradiacion parcial de mama con IMRT

    Energy Technology Data Exchange (ETDEWEB)

    Velazquez T, J. J.; Gutierrez M, J. G.; Ortiz A, C. S.; Chagoya G, A.; Gutierrez C, J. G., E-mail: jvelaesfm@gmail.com [Centro Medico Nacional Siglo XXI, Hospital de Oncologia, Departamentos de Fisica Medica y Radioterapia, Av. Cuauhtemoc 330, 03020 Mexico D. F. (Mexico)

    2015-10-15

    Partial breast irradiation is a new type of external radiation therapy to treat breast cancer in early clinical stages. Consist of administering to the channel surgical high doses of radiation in few treatment sessions. In this paper the dose calculations of the planning system Eclipse version 10 for a treatment of partial breast irradiation with X-rays beams (6 MV) intensity modulated were compared against the measurements made with OSL dosimeters and radio-chromic dye film. An anthropomorphic mannequin was used in which OSL dosimeters were collocated near the surface, an inside the radio-chromic dye film one plate; with this latest one dimensional dose distribution was measured. Previously dosimeters were calibrated irradiating them with a beam of X-rays 6 MV under the conditions specified in the IAEA-398 protocol. The OSL dosimeters were read in the Micro star Landauer equipment, the radio-chromic dye films were read with a scanner Epson 10000-Xl and analyzed with FilmCal and PTW Verisoft programs. The differences between measured and calculated dose were as follows: 3.6±1% for the OSL dosimeter and 96.3±1% of the analyzed points approved the gamma index criterion (3%, 3m m) when comparing the matrices of calculated dose and measured with the radio-chromic dye film. These results confirm the good dosimetric performance of planning system used under specific conditions used in the partial breast irradiation technique. (Author)

  7. Dosimetric validation of planning system Eclipse 10 in partial breast irradiation treatments with IMRT

    International Nuclear Information System (INIS)

    Velazquez T, J. J.; Gutierrez M, J. G.; Ortiz A, C. S.; Chagoya G, A.; Gutierrez C, J. G.

    2015-10-01

    Partial breast irradiation is a new type of external radiation therapy to treat breast cancer in early clinical stages. Consist of administering to the channel surgical high doses of radiation in few treatment sessions. In this paper the dose calculations of the planning system Eclipse version 10 for a treatment of partial breast irradiation with X-rays beams (6 MV) intensity modulated were compared against the measurements made with OSL dosimeters and radio-chromic dye film. An anthropomorphic mannequin was used in which OSL dosimeters were collocated near the surface, an inside the radio-chromic dye film one plate; with this latest one dimensional dose distribution was measured. Previously dosimeters were calibrated irradiating them with a beam of X-rays 6 MV under the conditions specified in the IAEA-398 protocol. The OSL dosimeters were read in the Micro star Landauer equipment, the radio-chromic dye films were read with a scanner Epson 10000-Xl and analyzed with FilmCal and PTW Verisoft programs. The differences between measured and calculated dose were as follows: 3.6±1% for the OSL dosimeter and 96.3±1% of the analyzed points approved the gamma index criterion (3%, 3m m) when comparing the matrices of calculated dose and measured with the radio-chromic dye film. These results confirm the good dosimetric performance of planning system used under specific conditions used in the partial breast irradiation technique. (Author)

  8. Establishing the impact of temporary tissue expanders on electron and photon beam dose distributions.

    Science.gov (United States)

    Asena, A; Kairn, T; Crowe, S B; Trapp, J V

    2015-05-01

    This study investigates the effects of temporary tissue expanders (TTEs) on the dose distributions in breast cancer radiotherapy treatments under a variety of conditions. Using EBT2 radiochromic film, both electron and photon beam dose distribution measurements were made for different phantoms, and beam geometries. This was done to establish a more comprehensive understanding of the implant's perturbation effects under a wider variety of conditions. The magnetic disk present in a tissue expander causes a dose reduction of approximately 20% in a photon tangent treatment and 56% in electron boost fields immediately downstream of the implant. The effects of the silicon elastomer are also much more apparent in an electron beam than a photon beam. Evidently, each component of the TTE attenuates the radiation beam to different degrees. This study has demonstrated that the accuracy of photon and electron treatments of post-mastectomy patients is influenced by the presence of a tissue expander for various beam orientations. The impact of TTEs on dose distributions establishes the importance of an accurately modelled high-density implant in the treatment planning system for post-mastectomy patients. Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  9. Temperature profile data from EBT casts in the Indian Ocean from 13 February 1986 to 01 January 1989 (NODC Accession 0000210)

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — Temperature profile data were collected using EBT casts in the Indian Ocean from the MYS OSTROVSKOGO, IGNAT PAVLYUCHENKOV, ZVEZDA AZOVA, and KARA-DAG from February...

  10. Effect of beta limits on reactor performance in EBT

    International Nuclear Information System (INIS)

    Uckan, N.A.; Spong, D.A.; Nelson, D.B.

    1981-01-01

    Because of uncertainties in extrapolating results of simplified models to a reactor plasma, the parameters that influence the beta limits cannot be determined accurately at the present time. Also, the reasonable changes within the models and/or assumptions are seen to affect the core beta limits by almost an order of magnitde. Hence, at the present, these limits cannot be used as a rigid (and reliable) requirement for ELMO Bumpy Torus (EBT) reactor engineering considerations. However, sensitivity studies can be carried out to determine the boundaries of the operating regime and to demonstrate the effects of various modes, assumptions, and models on reactor performance (Q value). First, the modes believed to limit the core β and ring plasma performance are discussed, and the simplifications and/or assumptions involved in deriving these limits are highlighted. Then, the implications of these limits for a reactor are given

  11. SU-E-T-104: Development of 3 Dimensional Dosimetry System for Gamma Knife

    Energy Technology Data Exchange (ETDEWEB)

    Yoon, K; Kwak, J; Cho, B; Lee, D; Ahn, S [Asan Medical Center, Seoul (Korea, Republic of)

    2014-06-01

    Purpose: The aim of this study was to develop a new 3 dimensional dosimetry system to verify the dosimetric accuracy of Leksell Gamma Knife-Perfexion™ (LGK) (Elekta, Norcross, GA). Methods: We designed and manufactured a lightweight dosimetry instrument to be equipped with the head frame to LGK. It consists of a head phantom, a scintillator, a CCD camera and a step motor. The 10×10 cm2 sheet of Gd2O3;Tb phosphor or Gafchromic EBT3 film was located at the center of the 16 cm diameter hemispherical PMMA, the head phantom. The additional backscatter compensating material of 1 cm thick PMMA plate was placed downstream of the phosphor sheet. The backscatter plate was transparent for scintillation lights to reach the CCD camera with 1200×1200 pixels by 5.2 um pitch. With This equipment, 300 images with 0.2 mm of slice gap were acquired under three collimator setups (4mm, 8mm and 16mm), respectively. The 2D projected doses from 3D distributions were compared with the exposured film dose. Results: As all doses normalized by the maximum dose value in 16 mm setup, the relative differences between the equipment dose and film dose were 0.2% for 4mm collimator and 0.5% for 8mm. The acquisition of 300 images by the equipment took less than 3 minutes. Conclusion: The new equipment was verified to be a good substitute to radiochromic film, with which required more time and resources. Especially, the new methods was considered to provide much convenient and faster solution in the 3D dose acquisition for LGK.

  12. Analysis of the Dose Distribution of Moving Organ using a Moving Phantom System

    International Nuclear Information System (INIS)

    Kim, Yon Lae; Park, Byung Moon; Bae, Yong Ki; Kang, Min Young; Bang, Dong Wan; Lee, Gui Won

    2006-01-01

    Few researches have been performed on the dose distribution of the moving organ for radiotherapy so far. In order to simulate the organ motion caused by respiratory function, multipurpose phantom and moving device was used and dosimetric measurements for dose distribution of the moving organs were conducted in this study. The purpose of our study was to evaluate how dose distributions are changed due to respiratory motion. A multipurpose phantom and a moving device were developed for the measurement of the dose distribution of the moving organ due to respiratory function. Acryl chosen design of the phantom was considered the most obvious choice for phantom material. For construction of the phantom, we used acryl and cork with density of 1.14 g/cm 3 , 0.32 g/cm 3 respectively. Acryl and cork slab in the phantom were used to simulate the normal organ and lung respectively. The moving phantom system was composed of moving device, moving control system, and acryl and cork phantom. Gafchromic film and EDR2 film were used to measure dose distributions. The moving device system may be driven by two directional step motors and able to perform 2 dimensional movements (x, z axis), but only 1 dimensional movement(z axis) was used for this study. Larger penumbra was shown in the cork phantom than in the acryl phantom. The dose profile and isodose curve of Gafchromic EBT film were not uniform since the film has small optical density responding to the dose. As the organ motion was increased, the blurrings in penumbra, flatness, and symmetry were increased. Most of measurements of dose distributions, Gafchromic EBT film has poor flatness and symmetry than EDR2 film, but both penumbra distributions were more or less comparable. The Gafchromic EBT film is more useful as it does not need development and more radiation dose could be exposed than EDR2 film without losing film characteristics. But as response of the optical density of Gafchromic EBT film to dose is low, beam profiles

  13. Evaluation of patients skin dose undergoing interventional cardiology procedure using radiochromic films

    International Nuclear Information System (INIS)

    Silva, Mauro W. Oliveira da; Canevaro, Lucia V.; Rodrigues, Barbara B. Dias

    2011-01-01

    In interventional cardiology (IC), coronary angiography (CA) and percutaneous transluminal coronary angioplasty (PTCA) procedures are the most frequent ones. Since the 1990s, the number of IC procedures has increased rapidly. It is also known that these procedures are associated with high radiation doses due to long fluoroscopy time (FT) and large number of cine-frames (CF) acquired to document the procedure. Mapping skin doses in IC is useful to find the probability of skin injuries, to detect areas of overlapping field, and to get a permanent record of the most exposed areas of skin. The purpose of this study was to estimate the maximum skin dose (MSD) in patients undergoing CA and PTCA, and to compare these values with the reference levels proposed in the literature. Patients' dose measurements were carried out on a sample of 38 patients at the hemodynamic department, in four local hospitals in Rio de Janeiro, Brazil, using Gafchromic XR-RV2 films. In PTCA procedures, the median and third quartile values of MSD were estimated at 2.5 and 5.3 Gy, respectively. For the CA procedures, the median and third quartile values of MSD were estimated at 0.5 and 0.7 Gy, respectively. In this paper, we used the Pearson's correlation coefficient (r), and we found a fairly strong correlation between FT and MSD (r=0.8334, p<0.0001), for CA procedures. The 1 Gy threshold for deterministic effects was exceeded in nine patients. The use of Gafchromic XR-RV2 films was shown to be an effective method to measure MSD and the dose distribution map. The method is effective to identify the distribution of radiation fields, thus allowing the follow-up of the patient to investigate the appearance of skin injuries. (author)

  14. Temperature and relative humidity dependence of radiochromic film dosimeter response to gamma electron radiation

    DEFF Research Database (Denmark)

    McLaughlin, W.L.; Puhl, J.M.; Miller, A.

    1995-01-01

    on some earlier studies, their response functions have been reported to be dependent on the temperature and relative humidity during irradiation. The present study investigates differences in response over practical ranges of temperature, relative humidity, dose, and for different recent batches of films...... humidity) and should be calibrated under environmental conditions (temperature) at which they will be used routinely....

  15. Dosimetric properties of a radiochromic gel detector for diagnostic X-rays

    International Nuclear Information System (INIS)

    Bero, M.A.

    2007-01-01

    The gel dosimetry method was found to be capable of addressing complicated issues related to dose measurements particularly in modern sophisticated radiotherapy applications. The Ferrous-sulphate Xylenol-orange and Gelatin (FXG) radiochromic gel dosemeter is one of the systems used for such applications. Some chemical dosemeters show different responses for low- and medium-energy X-rays in comparison with high-energy γ-photons. The energy and dose rate dependence of the FXG dose response was examined. In addition to the detector response, other important dosimetric properties of the system were investigated for different X-ray beam qualities with tube voltages in the range 100-300 kV. An orthovoltage X-ray therapy unit was used to irradiate standard sized samples of FXG from different batches for radiation doses in the range 0-20 Gy

  16. Patient skin dosimetry in interventional cardiology in the Czech Republic

    International Nuclear Information System (INIS)

    Sukupova, L.; Novak, L.; Kala, P.; Cervinka, P.; Stasek, J.

    2011-01-01

    In this study, skin dosimetry of patients undergoing interventional cardiology procedures is presented. Three hospitals were included. Two methods were used for skin dosimetry-radiochromic dosimetry films and reconstruction of skin dose distribution based on examination protocol. Maximum skin doses (MSD) obtained from both methods were compared for 175 patients. For patients for whom the film MSD was >1 Gy, the reconstruction MSD differed from the film MSD in the range of ± 50 % for 83 % of patients. For remaining patients, the difference was higher and it was caused by longer fluoroscopy time. For 59 patients for whom the cumulative dose was known, the cumulative dose was compared with the film MSD. Skin dosimetry with radiochromic films is more accurate than the reconstruction method, but films do not include X-ray fields from lateral projections whilst reconstructions do. (authors)

  17. EBT Payment for Online Grocery Orders: a Mixed-Methods Study to Understand Its Uptake among SNAP Recipients and the Barriers to and Motivators for Its Use.

    Science.gov (United States)

    Martinez, Olivia; Tagliaferro, Barbara; Rodriguez, Noemi; Athens, Jessica; Abrams, Courtney; Elbel, Brian

    2018-04-01

    To examine Supplemental Nutrition Assistance Program (SNAP) recipients' use of the first online supermarket accepting Electronic Benefit Transfer (EBT) payment. In this mixed-methods study, the authors collected EBT purchase data from an online grocer and attempted a randomized controlled trial in the South Bronx, New York City, followed by focus groups with SNAP beneficiaries aged ≥18 years. Participants were randomized to shop at their usual grocery store or an online supermarket for 3 months. Focus groups explored barriers and motivators to online EBT redemption. Few participants made online purchases, even when incentivized in the randomized controlled trial. Qualitative findings highlighted a lack of perceived control over the online food selection process as a key barrier to purchasing food online. Motivators included fast, free shipping and discounts. Electronic Benefit Transfer for online grocery purchases has the potential to increase food access among SNAP beneficiaries, but challenges exist to this new food buying option. Understanding online food shopping barriers and motivators is critical to the success of policies targeting the online expansion of SNAP benefits. Copyright © 2017 Society for Nutrition Education and Behavior. Published by Elsevier Inc. All rights reserved.

  18. Initial color development in radiochromic dye films after a short intense pulse of accelerated electrons

    International Nuclear Information System (INIS)

    Uribe, R.M.; Barcelo, M.; Rios, J.; McLaughlin, W.L.; Buenfil, A.E.

    1990-01-01

    The radiation response of different dye precursors in several host plastics has been investigated after a single short-pulse irradiation with 2.5-MeV electrons. It was observed that in most films the radiation-initiated color development proceeds mainly during the first 300 seconds, after such high dose-rate irradiation (∼ 10 12 Gy/s). Absorption spectra show that the main absorption band increases at the expense of a shorter-wavelength precursor absorption band, showing an isosbestic point approximately midway bwetwen the two absorption bands. It was found that a certain combination of dye precursor and host plastic (namely a polyamide containing an aromatic group) constitutes a film which shows a very fast increase in optical density of the main absorption band, making it suitable for immediate dosimetric analysis in very high dose-rate installations. (author)

  19. Radiochromic Plastic Films for Accurate Measurement of Radiation Absorbed Dose and Dose Distributions

    DEFF Research Database (Denmark)

    McLaughlin, W. L.; Miller, Arne; Fidan, S.

    1977-01-01

    of dose rate (1–1014 rad s−1). Upon irradiation of the film, the profile of the radiation field is registered as a permanent colored image of the dose distribution. Unlike most other types of dyed plastic dose meters, the optical density produced by irradiation is in most cases stable for periods...... of many polymeric systems in industrial radiation processing. The result is that errors due to energy dependence of response of the radiation sensor are effectively reduced, since the spectral sensitivity of the dose meter matches that of the polymer of interest, over a wide range of photon and electron...

  20. The radiochromic dye film dose meter as a possible test of particle track theory

    International Nuclear Information System (INIS)

    Hansen, J.W.; Jensen, M.; Katz, R.

    1980-09-01

    The response characteristic of the thin-film radiometric dye cyanide plastic dose meter to ionizing radiation of electrons and heavy charged particles is investigated as a possible test of the particle track theory worked out by Robert Katz and coworkers. Dose response curves for low-LET radiation have been investigated and are used for a quality estimation of the response for protons and oxygen ions at 16 and 4 MeV/amu, respectively. A bleaching effect on the colouration at high doses intimates that the target cannot be interpreted lieerally, but it might still be possinle to transfer the function of the macroscopic dose response to a theoretical dose response curve in a microscopic scale for a single ion. From this relation the macroscopic dose response curve can be determined qhen the film is irradiated with heavy ions. It will be shown theoretically that for protons there is no saturation in the track core, whereas calculations for oxygen ions show a heavy saturation in the track core, which means that a part of the ions loose their energy ineffectively. We can conclude that itis possible qualitatively to predict the dose response curve for high-LET particles by means of the dose response curve for low-LET radiation. (author)

  1. Spectrophotometric determination of β-adrenergic antagonists drugs via ion-pair complex formation using MO and EBT

    Science.gov (United States)

    El-Didamony, A. M.; Shehata, A. M.

    2014-09-01

    Two simple, rapid and sensitive spectrophotometric methods have been proposed for the assay of bisoprolol fumarate (BSF), propranolol hydrochloride (PRH), and timolol maleate (TIM) either in bulk or in pharmaceutical formulations. The methods are based on the reaction of the selected drugs with methyl orange (MO) and eriochrome black T in acidic buffers, after extracting in dichloromethane and measured quantitatively with maximum absorption at 428 and 518 nm for MO and EBT, respectively. The analytical parameters and their effects on the reported systems are investigated. The extracts are intensely colored and very stable at room temperature. The calibration graphs were linear over the concentration range of 0.8-6.4, 0.4-3.6, 0.8-5.6 μg/mL for BSF, PRH, and TIM, respectively, with MO and 0.8-6.4, 0.4-3.2, and 0.8-8.0 μg/mL for BSF, PRH, and TIM, respectively, with EBT. The stoichiometry of the complexes was found to be 1 : 1 in all cases. The proposed methods were successfully extended to pharmaceutical preparations. Excipients used as additive in commercial formulations did not interfere in the analysis. The proposed methods can be recommended for quality control and routine analysis where time, cost effectiveness and high specificity of analytical technique are of great importance.

  2. Irradiated polymeric films and its possible application for radiation dosimetry

    International Nuclear Information System (INIS)

    Elkady, A.A.M.A.

    2002-01-01

    The aim of the present thesis is the development of a new polymeric films, laboratory prepared by sample uncouthly methods through dying them with different types of dyes. First unit containing three forms were prepared by casting polymeric solution of poly (vinyl chloride). First film was prepared by dying poly ( vinyl chloride ) film with eosin dye. This flexible plastic film dosimeter is bleached when exposed to gamma ray photons at low doses (2-7 kGy) indicating their suitability for eventual application in food irradiation processing. Also, another polyvinyl chloride film contain a Ph indicating dye namely methyl red undergo bleaching in the rang of (20-120 kGy) indicating their suitability for eventual application in sterilization. The third film was a Ph indicating dye cresol red dyed polyvinyl chloride in the presence of CI- containing substance (E.g.Chloralhydrate). This film have the advantage of clear visual in color at high dose gamma radiation. This film was developed using the phenomenon of HCl generation from the CI- containing substance upon irradiation. The produced HCl in turn changes the color of the indicator contained in the polymer. The Dosimetric parameters e.g. dose response, effect of relative humidity and temperature during irradiation on response as well as pre-and post-irradiation stability of all above films are discussed. In second unit, the film was prepared by sandwiching the radiochromic film (nylon 66) containing (Hexa-hydroxyethyl pararosaniline) by two layers of polyvinyl chloride containing UV absorption material namely (2-hydroxy-4n-octoxy-benzophenone) to improve the stability of this famous film towards UV-radiation. Finally, a mixed dyes of ph indicating dyes bromophenol red and cresol red dying a poly vinyl alcohol in the presence of chloral hydrate were prepared. This film exhibit color changes from original color (bagel), yellow to red. The useful dose range of this is (5-50 kGy) indicate its suitability for eventual

  3. Film dosimetry using a smart device camera: a feasibility study for point dose measurements

    Science.gov (United States)

    Aland, Trent; Jhala, Ekta; Kairn, Tanya; Trapp, Jamie

    2017-10-01

    In this work, a methodology for using a smartphone camera, in conjunction with a light-tight box operating in reflective transmission mode, is investigated as a proof of concept for use as a film dosimetry system. An imaging system was designed to allow the camera of a smartphone to be used as a pseudo densitometer. Ten pieces of Gafchromic EBT3 film were irradiated to doses up to 16.89 Gy and used to evaluate the effects of reproducibility and orientation, as well as the ability to create an accurate dose response curve for the smartphone based dosimetry system, using all three colour channels. Results were compared to a flatbed scanner system. Overall uncertainty was found to be best for the red channel with an uncertainty of 2.4% identified for film irradiated to 2.5 Gy and digitised using the smartphone system. This proof of concept exercise showed that although uncertainties still exceed a flatbed scanner system, the smartphone system may be useful for providing point dose measurements in situations where conventional flatbed scanners (or other dosimetry systems) are unavailable or unaffordable.

  4. Film dosimetry using a smart device camera: a feasibility study for point dose measurements.

    Science.gov (United States)

    Aland, Trent; Jhala, Ekta; Kairn, Tanya; Trapp, Jamie

    2017-10-03

    In this work, a methodology for using a smartphone camera, in conjunction with a light-tight box operating in reflective transmission mode, is investigated as a proof of concept for use as a film dosimetry system. An imaging system was designed to allow the camera of a smartphone to be used as a pseudo densitometer. Ten pieces of Gafchromic EBT3 film were irradiated to doses up to 16.89 Gy and used to evaluate the effects of reproducibility and orientation, as well as the ability to create an accurate dose response curve for the smartphone based dosimetry system, using all three colour channels. Results were compared to a flatbed scanner system. Overall uncertainty was found to be best for the red channel with an uncertainty of 2.4% identified for film irradiated to 2.5 Gy and digitised using the smartphone system. This proof of concept exercise showed that although uncertainties still exceed a flatbed scanner system, the smartphone system may be useful for providing point dose measurements in situations where conventional flatbed scanners (or other dosimetry systems) are unavailable or unaffordable.

  5. Breast internal dose measurements in a physical thoracic phantom

    Energy Technology Data Exchange (ETDEWEB)

    Silva, S.D.; Oliveira, M.A.; Castro, A.L.S.; Dias, H.G.; Nogueira, L.B.; Campos, T.P.R., E-mail: sadonatosilva@hotmail.com [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Departamento de Engenharia Nuclear; Hospital das Clinicas de Uberlandia, MG (Brazil). Departamento de Oncologia; Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Anatomia e Departamento de Imagem

    2017-10-01

    Radiotherapy is a cancer treatment intended to deposit the entire prescribed dose homogeneously into a target volume in order to eliminate the tumor and to spare the surrounding healthy tissues. This paper aimed to provide a dosimetric comparison between the treatment planning system (TPS) ECLIPSE from Varian Medical Systems and the internal dosimetric measurements in a breast phantom. The methodology consisted in performing a 3D conformal radiotherapy planning with two tangential opposite parallel fields applied to the synthetic breast in a thoracic phantom. The irradiation was reproduced in the Varian Linear accelerator, model SL - 20 Precise, 6 MV energy. EBT2 Radiochromic films, placed into the glandular equivalent tissue of the breast, were used to measure the spatial dose distribution. The absorbed dose was compared to those values predicted by the treatment planning system; besides, the dosimetric uncertainties were analyzed. The modal absorbed dose was in agreement with the prescribed value of 180 cGy, although few high dose points between 180 and 220 cGy were detected. The findings suggested a non-uniform dose distribution in the glandular tissue of the synthetic breast, similar to those found in the TPS, associated with the irregular anatomic breast shape and presence of inhomogeneities next to the thoracic wall generated by the low lung density. (author)

  6. Technical Note: Response time evolution of XR-QA2 GafChromic™ film models.

    Science.gov (United States)

    Aldelaijan, Saad; Tomic, Nada; Papaconstadopoulos, Pavlos; Schneider, James; Seuntjens, Jan; Shih, Shelley; Lewis, David; Devic, Slobodan

    2018-01-01

    To evaluate the response of the newest XR-QA2 GafChromic™ film model in terms of postexposure signal growth and energy response in comparison with the older XR-QA (Version 2) model. Pieces of film were irradiated to air kerma in air values up to 12 cGy with several beam qualities (5.3-8.25 mm Al) commonly used for CT scanning. Film response was scored in terms of net reflectance from scanned film images at various points in time postirradiation ranging from 1 to 7 days and 5 months postexposure. To reconstruct the measurement signal changes with postirradiation delay, we irradiated one film piece and then scanned it at different point times starting from 2" min and up to 3 days postexposure. For all beam qualities and dose range investigated, it appears that the XR-QA2 film signal completely saturated after 15 h. Compared to 15 h postirradiation scanning time, the observed variation in net reflectance were 3%, 2%, and 1% for film scanned 2" min, 20 min, and 3 h after exposure, respectively, which is well within the measurement uncertainty of the XR-QA2 based reference radiochromic film dosimetry system. A comparison between the XR-QA (Version 2) and the XR-QA2 film response after several months (relative to their responses after 24 h) show differences in up to 8% and 1% for each film model respectively. The replacement of cesium bromide in the older XR-QA (Version 2) film model with bismuth oxide in the newer XR-QA2 film, while keeping the same single sensitive layer structure, lead to a significantly more stable postexposure response. © 2017 American Association of Physicists in Medicine.

  7. SU-C-201-02: Dosimetric Verification of SBRT with FFF-VMAT Using a 3-D Radiochromic/Optical-CT Dosimetry System

    Energy Technology Data Exchange (ETDEWEB)

    Na, Y; Black, P; Wuu, C [Columbia University, New York, NY (United States); Adamovics, J [Department of Chemistry and Biology, Rider University, Skillman, NJ (United States)

    2016-06-15

    Purpose: With an increasing use of small field size and high dose rate irradiation in the advances of radiotherapy techniques, such as stereotactic body radiotherapy (SBRT) and stereotactic radiosurgery (SRS), an in-depth quality assurance (QA) system is required. The purpose of this study is to investigate a high resolution optical CT-based 3D radiochromic dosimetry system for SBRT with intensity modulated radiotherapy (IMRT) and flattening filter free (FFF) volumetric modulated arc therapy (VMAT). Methods: Cylindrical PRESAGE radiochromic dosimeters of 10cm height and 11cm diameter were used to validate SBRT. Four external landmarks were placed on the surface of each dosimeter to define the isocenter of target. SBRT plans were delivered using a Varian TrueBeam™ linear accelerator (LINAC). Three validation plans, SBRT with IMRT (6MV 600MU/min), FFF-VMAT (10MV 2400MU/min), and mixed FFF-VMAT (6MV 1400MU/min, 10MV 2400MU/min), were delivered to the PRESAGE dosimeters. Each irradiated PRESAGE dosimeter was scanned using a single laser beam optical CT scanner and reconstructed with a 1mm × 1mm high spatial resolution. The comparison of measured dose distributions of irradiated PRESAGE dosimeters to those calculated by Pinnacle{sup 3} treatment planning system (TPS) were performed with a 10% dose threshold, 3% dose difference (DD), and 3mm distance-to-agreement (DTA) Gamma criteria. Results: The average pass rates for the gamma comparisons between PRESAGE and Pinnacle{sup 3} in the transverse, sagittal, coronal planes were 94.6%, 95.9%, and 96.4% for SBRT with IMRT, FFF-VMAT, and mixed FFF-VMAT plans, respectively. A good agreement of the isodose distributions of those comparisons were shown at the isodose lines 50%, 70%, 80%, 90% and 98%. Conclusion: This study demonstrates the feasibility of the high resolution optical CT-based 3D radiochromic dosimetry system for validation of SBRT with IMRT and FFF-VMAT. This dosimetry system offers higher precision QA with 3D

  8. Skin dose for head and neck cancer patients treated with intensity-modulated radiation therapy(IMRT)

    Science.gov (United States)

    Fu, Hsiao-Ju; Li, Chi-Wei; Tsai, Wei-Ta; Chang, Chih-Chia; Tsang, Yuk-Wah

    2017-11-01

    The reliability of thermoluminescent dosimeters (ultrathin TLD) and ISP Gafchromic EBT2 film to measure the surface dose in phantom and the skin dose in head-and-neck patients treated with intensity-modulated radiation therapy technique(IMRT) is the research focus. Seven-field treatment plans with prescribed dose of 180 cGy were performed on Eclipse treatment planning system which utilized pencil beam calculation algorithm(PBC). In calibration tests, the variance coefficient of the ultrathin TLDs were within 3%. The points on the calibration curve of the Gafchromic film was within 1% variation. Five measurements were taken on phantom using ultrathin TLD and EBT2 film respectively. The measured mean surface doses between ultrathin TLD or EBT2 film were within 5% deviation. Skin doses of 6 patients were measured for initial 5 fractions and the mean dose per-fraction was calculated. If the extrapolated doses for 30 fractions were below 4000 cGy, the skin reaction grading observed according to Radiation Therapy Oncology Group (RTOG) was either grade 1 or grade 2. If surface dose exceeded 5000 cGy in 32 fractions, then grade 3 skin reactions were observed.

  9. Dosimetry auditing procedure with alanine dosimeters for light ion beam therapy

    International Nuclear Information System (INIS)

    Ableitinger, Alexander; Vatnitsky, Stanislav; Herrmann, Rochus; Bassler, Niels; Palmans, Hugo; Sharpe, Peter; Ecker, Swantje; Chaudhri, Naved; Jäkel, Oliver; Georg, Dietmar

    2013-01-01

    Background and purpose: In the next few years the number of facilities providing ion beam therapy with scanning beams will increase. An auditing process based on an end-to-end test (including CT imaging, planning and dose delivery) could help new ion therapy centres to validate their entire logistic chain of radiation delivery. An end-to-end procedure was designed and tested in both scanned proton and carbon ion beams, which may also serve as a dosimetric credentialing procedure for clinical trials in the future. The developed procedure is focused only on physical dose delivery and the validation of the biological dose is out of scope of the current work. Materials and methods: The audit procedure was based on a homogeneous phantom that mimics the dimension of a head (20 × 20 × 21 cm 3 ). The phantom can be loaded either with an ionisation chamber or 20 alanine dosimeters plus 2 radiochromic EBT films. Dose verification aimed at measuring a dose of 10 Gy homogeneously delivered to a virtual-target volume of 8 × 8 × 12 cm 3 . In order to interpret the readout of the irradiated alanine dosimeters additional Monte Carlo simulations were performed to calculate the energy dependent detector response of the particle fluence in the alanine detector. A pilot run was performed with protons and carbon ions at the Heidelberg Ion Therapy facility (HIT). Results: The mean difference of the absolute physical dose measured with the alanine dosimeters compared with the expected dose from the treatment planning system was −2.4 ± 0.9% (1σ) for protons and −2.2 ± 1.1% (1σ) for carbon ions. The measurements performed with the ionisation chamber indicate this slight underdosage with a dose difference of −1.7% for protons and −1.0% for carbon ions. The profiles measured by radiochromic films showed an acceptable homogeneity of about 3%. Conclusions: Alanine dosimeters are suitable detectors for dosimetry audits in ion beam therapy and the presented end-to-end test is

  10. Dosimetry auditing procedure with alanine dosimeters for light ion beam therapy.

    Science.gov (United States)

    Ableitinger, Alexander; Vatnitsky, Stanislav; Herrmann, Rochus; Bassler, Niels; Palmans, Hugo; Sharpe, Peter; Ecker, Swantje; Chaudhri, Naved; Jäkel, Oliver; Georg, Dietmar

    2013-07-01

    In the next few years the number of facilities providing ion beam therapy with scanning beams will increase. An auditing process based on an end-to-end test (including CT imaging, planning and dose delivery) could help new ion therapy centres to validate their entire logistic chain of radiation delivery. An end-to-end procedure was designed and tested in both scanned proton and carbon ion beams, which may also serve as a dosimetric credentialing procedure for clinical trials in the future. The developed procedure is focused only on physical dose delivery and the validation of the biological dose is out of scope of the current work. The audit procedure was based on a homogeneous phantom that mimics the dimension of a head (20 × 20 × 21 cm(3)). The phantom can be loaded either with an ionisation chamber or 20 alanine dosimeters plus 2 radiochromic EBT films. Dose verification aimed at measuring a dose of 10Gy homogeneously delivered to a virtual-target volume of 8 × 8 × 12 cm(3). In order to interpret the readout of the irradiated alanine dosimeters additional Monte Carlo simulations were performed to calculate the energy dependent detector response of the particle fluence in the alanine detector. A pilot run was performed with protons and carbon ions at the Heidelberg Ion Therapy facility (HIT). The mean difference of the absolute physical dose measured with the alanine dosimeters compared with the expected dose from the treatment planning system was -2.4 ± 0.9% (1σ) for protons and -2.2 ± 1.1% (1σ) for carbon ions. The measurements performed with the ionisation chamber indicate this slight underdosage with a dose difference of -1.7% for protons and -1.0% for carbon ions. The profiles measured by radiochromic films showed an acceptable homogeneity of about 3%. Alanine dosimeters are suitable detectors for dosimetry audits in ion beam therapy and the presented end-to-end test is feasible. If further studies show similar results, this dosimetric audit could be

  11. Wall effects on the absorption of electron cyclotron waves in an EBT plasma

    International Nuclear Information System (INIS)

    Uckan, T.

    1979-03-01

    The absorption of electron cyclotron waves propagating along an externally applied magnetic field in a uniform plasma surrounded by a cylindrical metallic cavity wall is studied. In the model, the cavity wall, the vacuum-plasma interface, and the effects of finite electron temperature are considered, and the dispersion relation for the wave propagation is derived. The results are then applied to the ELMO Bumpy Torus (EBT-I) plasma, and the propagation characteristics are computed. The wave absorption in the ordinary mode is found to be a result of the wall effects, which cannot be predicted with the infinite plasma theory. The loaded quality factor, Q/sub L/, is also estimated from the model to be about 12, which is in good agreement with the experimentally observed value

  12. A method of dosimetry for synchrotron microbeam radiation therapy using radiochromic films and microdensitometry

    International Nuclear Information System (INIS)

    Crosbie, J. C.; Svalbe, I. D.; Lewis, R. A.

    2007-01-01

    Full text: Normal tissue displays an exceptional tolerance to high doses of radiation (hundreds of Gy) when delivered as a microplanar array of synchrotron-generated x-rays. Furthermore, MRT has been shown to cause significant tumour growth delay and in some case complete ablation. The biological effects of MRT on tissue are not fully understood. This is further complicated by difficulties in performing accurate dosimetry. The majority of dosimetry performed for MRT has been Monte Carlo simulations. The aim of this work was to utilise film dosimetry and microdensitometry to measure the peak-to-valley dose ratios (PVDRs) for synchrotron microbeam radiation therapy.

  13. Curvature-driven instabilities in the Elmo Bumpy Torus (EBT)

    International Nuclear Information System (INIS)

    Abe, H.; Spong, D.A.; Antonsen, T.M. Jr.; Tsang, K.T.; Nguyen, K.T.

    1982-01-01

    Curvature-driven instabilities are analyzed for an EBT configuration which consists of plasma interacting with a hot electron ring whose drift frequencies are larger than the growth rates predicted from conventional magnetohydrodynamic (MHD) theory. Stability criteria are obtained for five possible modes: the conventional hot electron interchange, a high-frequency hot electron interchange (at frequencies greater than the ion-cyclotron frequency), a compressional instability, a background plasma interchange, and an interacting pressure-driven interchange. A wide parameter regime for stable operation is found, which, however, severely deteriorates for a band of intermediate mode numbers. Finite Larmor radius effects can eliminate this deterioration; moreover, all short-wavelength curvature-driven modes are stabilized if the hot electron Larmor radius rho/sub h/ satisfies (kappa/sub perpendicular/rho/sub h/) 2 > 2Δ/[Rβ/sub h/(1 + P'/sub parallel//P'/sub perpendicular/)], where kappa/sub perpendicular/ is the transverse wavenumber, Δ is the ring half-width, R is the mid-plane radius of curvature, β/sub h/ is the hot electron beta value, and P' is the pressure gradient. Resonant wave-particle instabilities predicted by a new low frequency variational principle show that a variety of remnant instabilities may still persist

  14. Radiological response and dosimetry in physical phantom of head and neck for 3D conformational radiotherapy; Resposta radiologica e dosimetria em phantom fisico de cabeca e pescoco para radioterapia conformacional 3D

    Energy Technology Data Exchange (ETDEWEB)

    Thompson, Larissa

    2013-07-01

    Phantoms are tools for simulation of organs and tissues of the human body in radiology and radiotherapy. This thesis describes the development, validation and, most importantly, the use of a physical head and neck phantom in radiology and radiotherapy, with the purpose of evaluating dose distribution using Gafchromic EBT2 film in 15 MV 3D conformal radiotherapy. The work was divided in two stages, (1) development of new equivalent tissues and improvement of the physical phantom, and (2) use of the physical phantom in experimental dosimetry studies. In phase (1) parameters such as mass density, chemical composition of tissues, anatomical and biometric measurements were considered, as well as aspects of imaging by computed tomography (CT) and radiological response representation in Hounsfield Units (HU), which were compared with human data. Radiological experiments of in-phantom simulated brain pathologies were also conducted. All those results matched human-sourced data, therefore the physical phantom is a suitable simulator that may be used to enhance radiological protocols and education in medical imaging. The main objective in phase (2) was to evaluate the spatial dose distribution in a brain tumor simulator inserted inside the head and neck phantom developed by the Ionizing Radiation Research Group (NRI), exposed to 15 MV 3D conformal radiotherapy, for internal dose assessment. Radiation planning was based on CT images of the physical phantom with a brain tumor simulator made with equivalent material. The treatment planning system (TPS), CAT3D software, used CT images and prescribed a dose of 200 cGy, distributed in three fields of radiation, in a T-shaped pattern. The TPS covered the planning treatment volume (PTV) with 97% of the prescribed dose. A solid water phantom and radiochromic Gafchromic EBT2 film were used for calibration procedures, generating a dose response curve as a function of optical density (OD). After calibration and irradiation, the film

  15. Radiological response and dosimetry in physical phantom of head and neck for 3D conformational radiotherapy

    International Nuclear Information System (INIS)

    Thompson, Larissa

    2013-01-01

    Phantoms are tools for simulation of organs and tissues of the human body in radiology and radiotherapy. This thesis describes the development, validation and, most importantly, the use of a physical head and neck phantom in radiology and radiotherapy, with the purpose of evaluating dose distribution using Gafchromic EBT2 film in 15 MV 3D conformal radiotherapy. The work was divided in two stages, (1) development of new equivalent tissues and improvement of the physical phantom, and (2) use of the physical phantom in experimental dosimetry studies. In phase (1) parameters such as mass density, chemical composition of tissues, anatomical and biometric measurements were considered, as well as aspects of imaging by computed tomography (CT) and radiological response representation in Hounsfield Units (HU), which were compared with human data. Radiological experiments of in-phantom simulated brain pathologies were also conducted. All those results matched human-sourced data, therefore the physical phantom is a suitable simulator that may be used to enhance radiological protocols and education in medical imaging. The main objective in phase (2) was to evaluate the spatial dose distribution in a brain tumor simulator inserted inside the head and neck phantom developed by the Ionizing Radiation Research Group (NRI), exposed to 15 MV 3D conformal radiotherapy, for internal dose assessment. Radiation planning was based on CT images of the physical phantom with a brain tumor simulator made with equivalent material. The treatment planning system (TPS), CAT3D software, used CT images and prescribed a dose of 200 cGy, distributed in three fields of radiation, in a T-shaped pattern. The TPS covered the planning treatment volume (PTV) with 97% of the prescribed dose. A solid water phantom and radiochromic Gafchromic EBT2 film were used for calibration procedures, generating a dose response curve as a function of optical density (OD). After calibration and irradiation, the film

  16. Dosimetric evaluation of a novel high dose rate (HDR) intraluminal / interstitial brachytherapy applicator for gastrointestinal and bladder cancers

    Science.gov (United States)

    Aghamiri, Seyyed Mahmoud Reza; Najarian, Siamak; Jaberi, Ramin

    2010-01-01

    High dose rate (HDR) brachytherapy is one of the accepted treatment modalities in gastro‐intestinal tract and bladder carcinomas. Considering the shortcoming of contact brachytherapy routinely used in gastrointestinal tract in treatment of big tumors or invasive method of bladder treatment, an intraluminal applicator with the capability of insertion into the tumor depth seems to be useful. This study presents some dosimetric evaluations to introduce this applicator to the clinical use. The radiation attenuation characteristics of the applicator were evaluated by means of two dosimetric methods including well‐type chamber and radiochromic film. The proposed 110 cm long applicator has a flexible structure made of stainless steel for easy passage through lumens and a needle tip to drill into big tumors. The 2 mm diameter of the applicator is thick enough for source transition, while easy passage through any narrow lumen such as endoscope or cystoscope working channel is ensured. Well‐chamber results showed an acceptably low attenuation of this steel springy applicator. Performing absolute dosimetry resulted in a correlation coefficient of R=0.9916(p‐value≈10−7) between standard interstitial applicator and the one proposed in this article. This study not only introduces a novel applicator with acceptable attenuation but also proves the response independency of the GAFCHROMIC EBT films to energy. By applying the dose response of the applicator in the treatment planning software, it can be used as a new intraluminal / interstitial applicator. PACS number: 87.53.Bn, 87.53.Jw, 29.40.Cs

  17. Sci-Sat AM: Radiation Dosimetry and Practical Therapy Solutions - 04: On 3D Fabrication of Phantoms and Experimental Verification of Patient Dose Computation Algorithms

    Energy Technology Data Exchange (ETDEWEB)

    Khan, Rao; Zavan, Rodolfo; McGeachy, Philip; Madamesila, Joseph; Villarreal-Barajas, Jose Eduardo [Radiation Oncology, Washington University School of Medicine, St Louis, MO, USA, Instituto de Física de São Carlos, Universidade de São Paulo, SP, Brazil, CancerCare Manitoba, Winnipeg,Manitoba, Physics and Astronomy, University of Calgary, Calgary, Alberta (Canada)

    2016-08-15

    Purpose: Transport based dose calculation algorithm Acuros XB (AXB) has been shown to accurately account for heterogeneities mostly through comparisons with Monte Carlo simulations. This study aims at providing additional experimental verification for AXB for flattened and unflattened clinical energies in low density phantoms of the same material. Materials and Methods: Polystyrene slabs were created using a bench-top 3D printer. Six slabs were printed at varying densities from 0.23 g/cm{sup 3} to 0.68 g/cm{sup 3}, corresponding to different density humanoid tissues. The slabs were used to form different single and multilayer geometries. Dose was calculated with AXB 11.0.31 for 6MV, 15MV flattened and 6FFF (flattening filter free) energies for field sizes of 2×2 cm{sup 2} and 5×5 cm{sup 2}. The phantoms containing radiochromic EBT3 films were irradiated. Absolute dose profiles and 2D gamma analyses were performed for 96 dose planes. Results: For all single slab, multislab configurations and energies, absolute dose differences between the AXB calculation and film measurements remained <3% for both fields, with slightly poor disagreement in penumbra. The gamma index at 2% / 2mm averaged 98% in all combinations of fields, phantoms and photon energies. Conclusions: The transport based dose algorithm AXB is in good agreement with the experimental measurements for small field sizes using 6MV, 6FFF and 15MV beams adjacent to low density heterogeneous media. This work provides sufficient experimental ground to support the use of AXB for heterogeneous dose calculation purposes.

  18. SU-E-T-296: Dosimetric Analysis of Small Animal Image-Guided Irradiator Using High Resolution Optical CT Imaging of 3D Dosimeters

    International Nuclear Information System (INIS)

    Na, Y; Qian, X; Wuu, C; Adamovics, J

    2015-01-01

    Purpose: To verify the dosimetric characteristics of a small animal image-guided irradiator using a high-resolution of optical CT imaging of 3D dosimeters. Methods: PRESAEGE 3D dosimeters were used to determine dosimetric characteristics of a small animal image-guided irradiator and compared with EBT2 films. Cylindrical PRESAGE dosimeters with 7cm height and 6cm diameter were placed along the central axis of the beam. The films were positioned between 6×6cm 2 cubed plastic water phantoms perpendicular to the beam direction with multiple depths. PRESAGE dosimeters and EBT2 films were then irradiated with the irradiator beams at 220kVp and 13mA. Each of irradiated PRESAGE dosimeters named PA1, PA2, PB1, and PB2, was independently scanned using a high-resolution single laser beam optical CT scanner. The transverse images were reconstructed with a 0.1mm high-resolution pixel. A commercial Epson Expression 10000XL flatbed scanner was used for readout of irradiated EBT2 films at a 0.4mm pixel resolution. PDD curves and beam profiles were measured for the irradiated PRESAGE dosimeters and EBT2 films. Results: The PDD agreements between the irradiated PRESAGE dosimeter PA1, PA2, PB1, PB2 and the EB2 films were 1.7, 2.3, 1.9, and 1.9% for the multiple depths at 1, 5, 10, 15, 20, 30, 40 and 50mm, respectively. The FWHM measurements for each PRESAEGE dosimeter and film agreed with 0.5, 1.1, 0.4, and 1.7%, respectively, at 30mm depth. Both PDD and FWHM measurements for the PRESAGE dosimeters and the films agreed overall within 2%. The 20%–80% penumbral widths of each PRESAGE dosimeter and the film at a given depth were respectively found to be 0.97, 0.91, 0.79, 0.88, and 0.37mm. Conclusion: Dosimetric characteristics of a small animal image-guided irradiator have been demonstrated with the measurements of PRESAGE dosimeter and EB2 film. With the high resolution and accuracy obtained from this 3D dosimetry system, precise targeting small animal irradiation can be achieved

  19. Poster — Thur Eve — 53: Novel Technique for the Measurement of Ultra-Superficial Doses Using Gafchromic Film

    Energy Technology Data Exchange (ETDEWEB)

    Marcos, M.; Devic, S. [Jewish General Hospital — McGill University Teaching Hospital, Montreal, QC (Canada)

    2014-08-15

    Purpose: Dose build-up and electron contamination are two closely related quantities with important implications in radiotherapy, yet they are quite difficult to measure with great certainty. We present a novel technique for measuring ultra-superficial doses. Method and Materials: We used Gafchromic EBT-3 film which have an effective point of measurement of roughly 153 micros (effective depth in water). By peeling off one of the polyester layers, the active layer becomes the top layer and we obtain a film with a effective point of measurement of 15 microns (effective depth in water). A film calibration was performed using a 180 kVp orthovoltage beam. Since the active layer of the film may have been compressed or perturbed during the peeling of clear polyester we use a triple-channel film calibration technique to minimize the effects of non-uniformity in the active layer. We measured surface doses of orthovoltage beams with lead cutouts in place to introduce contaminant photoelectrons. Results: Our measurements show that the dose enhancement near the edges of the lead were about 125% relative to central axis for 6 cm diameter cutouts up to 170% for 2 cm diameter cutouts, which were within 5% of our EGSnrc based Monte Carlo simulations.

  20. The role of fusion reaction products in the stability of EBT reactors

    International Nuclear Information System (INIS)

    Wojtowicz, D.; Kammash, T.

    1985-01-01

    The potential of the EBT plasma confinement device as a fusion reactor depends critically on its ability to support a sufficiently large power density which in turn means a large enough beta defined as the ratio of the plasma pressure to magnetic field pressure. The maximum allowable beta is generally dictated by the stability of the system to hydromagnetic (MHD) modes. In this paper we examine the stability of such modes for a D-T plasma and assess the effect of the alpha particles on these instabilities. We find that the alphas have the most destabilizing effect, as reflected in the drop of the ion beta, at the instant of birth and that recovery of stability is achieved as the alphas approach equilibration with the ions of the plasma. In short, there appears to be no serious adverse effects on the reactor beta resulting from alpha-induced instabilities